United States
Environmental Protection
Agencv
Office of Air Quality
Planning and Standards
Research Triangle Park NC 27711
EMB Report 84-GLS-9
October 1984
Air
NESHAP - Glass
Manufacturing
Arsenic

Emission Test Report
Indiana Glass Company
Dunkirk, Indiana

-------
             SUMMARY TEST REPORT

      STANDARDIZATION AND VALIDATION OF
      METHODOLOGY TO MEASURE INORGANIC
           ARSENIC EMISSIONS FROM
             STATIONARY SOURCES
            Indiana Glass Company
              Dunkirk, Indiana

                     by

            PEI Associates, Inc.
    (formerly PEDCo Environmental, Inc.)
             11499 Chester Road
               P.O. Box 46100
        Cincinnati, Ohio  45246-0100
           Contract No. 68-02-3767
           Work Assignment No. 66
                Change No. 1
                  PN 3583-6
                     and
           Contract No. 68-02-3849
           Work Assignment No. 11
                 PN 3615-11
                Task Manager
              Mr. Daniel Bivins
 Emission Standards and Engineering Division
         Emission Measurement Branch
    U.S. ENVIRONMENTAL PROTECTION AGENCY
RESEARCH TRIANGLE PARK, NORTH CAROLINA  27711

                October 1984

-------
                           DISCLAIMER
     This report was furnished to the U.S. Environmental Protec-
tion Agency, Emission Measurement Branch, by PEI Associates,
Inc., Cincinnati, Ohio, in fulfillment of Contract No.  68-02-
3767, Work Assignment No. 66, Change No. 1.  Its contents are
reproduced herein as received from PEI.  The opinions,  findings,
and conclusions are those of the authors and not necessarily
those of the EPA.  Mention of company or product names  does not
constitute endorsement or recommendation for use.
                                11

-------
                            CONTENTS

                                                            Paqe
Figures                                                     iv
Tables                                                      vi
Acknowledgment                                              vii

1.   Introduction                                           1-1

2.   Summary and Discussion of Test Results                 2-1

     2.1  Sampling and analytical protocol                  2-1
     2.2  Test results—elevated temperature runs           2-2
     2.3  Method 108 traverse test results                  2-15
     2.4  Process samples                                   2-18

3.   Project Quality Assurance                              3-1

4.   Sampling Location and Test Methods                     4-1

     4.1  Sampling and analytical procedures                4-7

5.   Process Operation                                      5-1

Appendices

A    Computer Printouts and Example Calculations            A-l
B    Field Data                                             B-l
C    Laboratory Data                                        C-l
D    Sampling and Analytical Procedures                     D-l
E    Equipment Calibration Procedures and Results           E-l
F    Process Description and Operation During Testing       F-l
                                111

-------
                             FIGURES

Number                                                      Page
 2-1      Quad Train System for Elevated Temperature Tests  2-3

 3-1      Pre-Test Audit Report:  Dry Gas Meter by Critical
            Orifice  (Meter Box FB-1, Train A)               3-6

 3-2      Pre-Test Audit Report:  Dry Gas Meter by Critical
            Orifice  (Meter Box FB-5, Train B)               3-7

 3-3      Pre-Test Audit Report:  Dry Gas Meter by Critical
            Orifice  (Meter Box FB-8, Train D)               3-8

 3-4      Pre-Test Audit Report:  Dry Gas Meter by Critical
            Orifice  (Meter Box FB-10, Train C)              3-9

 3-5      Pre-Test Audit Report:  Dry Gas Meter by Critical
            Orifice  (Meter Box FB-9, Single Point-Traverse
            Tests)                                          3-10

 3-6      Pre-Test Thermocouple Digital Indicator Audit
            Data Sheet  (Indicator No. 220)                  3-11

 3-7      Pre-Test Thermocouple Digital Indicator Audit
            Data Sheet  (Indicator No. 221)                  3-12

 3-8      Pre-Test Onsite Audit Data Sheet                  3-13

 3-9      Pre-Test Onsite Audit Data Sheet                  3-14

 3-10     Mid-Test Audit Report:  Dry Gas Meter by Critical
            Orifice  (Meter Box FB-1)                        3-15

 3-11     Mid-Test Audit Report:  Dry Gas Meter by Critical
            Orifice  (Meter Box FB-5)                        3-16

 3-12     Mid-Test Audit Report:  Dry Gas Meter by Critical
            Orifice  (Meter Box FB-8)                        3-17

 3-13     Mid-Test Audit Report:  Dry Gas Meter by Critical
            Orifice  (Meter Box FB-10)                       3-18
                               IV

-------
                       FIGURES (continued)

Number                                                      Page

 3-14     Mid-Test Thermocouple Digital Indicator Audit Data
            Sheet (Indicator No.  220)                       3-19

 3-15     Mid-Test Thermocouple Digital Indicator Audit Data
            Sheet (Indicator No.  221)                       3-20

 3-16     Mid-Test Onsite Audit Data Sheet                  3-21

 3-17     Mid-Test Onsite Audit Data Sheet                  3-22

 3-18     Post-Test Audit Report:  Dry Gas Meter by Critical
            Orifice (Meter Box FB-1)                        3-23

 3-19     Post-Test Audit Report:  Dry Gas Meter by Critical
            Orifice (Meter Box FB-5)                        3-24

 3-20     Post-Test Audit Report:  Dry Gas Meter by Critical
            Orifice (Meter Box FB-8)                        3-25

 3-21     Post-Test Audit Report:  Dry Gas Meter by Critical
            Orifice (Meter Box FB-11)                       3-26

 3-22     Post-Test Thermocouple Digital Indicator Audit Data
            Sheet (Indicator No.  220)                       3-27

 3-23     Post-Test Thermocouple Digital Indicator Audit Data
            Sheet (Indicator No.  221)                       3-28

 3-24     Example of Unacceptable Dry Gas Meter Audit       3-29

 3-25     Example of Onsite Calibration Data Sheet          3-31

 4-1      Quad Train System for Elevated Temperature Tests  4-2

 4-2      Four-Train Sampling System Showing Nozzle, Pitot
            Tube, and Thermocouple Position                 4-3

 4-3      Furnace Exit Stack Elevation                      4-5

 4-4      Furnace Exit Stack Sampling Port Location         4-6
                               v

-------
                             TABLES

Number                                                      Page

 2-1      Summary of Sample Conditions                      2-4

 2-2      Summary of Arsenic Analytical Results - Quad
            and Reference Train Runs                        2-5

 2-3      Statistical Data for Grouped Runs                 2-7

 2-4      Statistical Data for Grouped Runs - EMSL Quad
            Train Tests                                     2-10

 2-5      Summary of Sample and Flue Gas Conditions
            Arsenic Traverse Tests                          2-16

 2-6      Summary of Arsenic Analytical Results - Traverse
            Train                                           2-17

 2-7      Process Sample Analytical Results                 2-19

 3-1      Field Equipment Calibration                       3-3

 3-2      Arsenic Blank Data                                3-32

 3-3      Arsenic Laboratory Reagent Blank Data             3-34

 3-4      Linear Regression Data  (Flame)                    3-36

 3-5      Arsenic Audit Results                             3-37

 3-6      Arsenic Standard Addition Results                 3-39

 3-7      Linear Regression Data  (Furnace)                  3-41

 3-8      Duplicate Analysis Data                           3-43
                               vi

-------
                         AC KNOWLEDGMENT






     This test program was conducted for the Emission Standards



and Engineering Division of the EPA Office of Air Quality Plan-



ning and Standards.  The program was part of a larger study



directed by EPA's Source Branch of the Environmental Monitoring



Systems Laboratory, Quality Assurance Division, to evaluate



proposed EPA Method 108.  Mr. Thomas E. Ward was the EMSL Task



Manager.



     Mr. Daniel Bivins, EPA-EMB Task Manager, provided overall



project coordination and guidance and observed the test program.



Mr. Ronald Myers, EPA lead engineer, Industrial Studies Branch,



provided project coordination relative to process operation and



overall project scope.  Mr. Larry Keller, representing Radian



Corporation (an EPA contractor) monitored process operation



throughout the test period.  Mr. Charles Bruffey was the PEI



Project Manager.  Principal authors were Messrs. Charles Bruffey



and Thomas Wagner.
                               VII

-------
                            SECTION 1
                          INTRODUCTION

     Arsenic is listed as a hazardous air pollutant under Section
112 of the Clean Air Act (National Emission Standards for Hazard-
ous Air Pollutants).  To protect public health from unreasonable
risks associated with exposure to airborne arsenic, the U.S.
Environmental Protection Agency (EPA) has developed standards to
decrease inorganic arsenic emissions from the following source
categories:  high-arsenic primary copper smelters, low-arsenic
primary copper smelters, and glass manufacturing plants.
     To support the standards review process and provide addi-
tional arsenic emissions data from glass manufacturing facili-

ties, PEI Associates, Inc., under contract to Research Triangle
Institute and directed by the Source Branch of the EMSL Quality
Assurance Division and the Emission Standards and Engineering
Division - Emission Measurement Branch, performed a series of
atmospheric emission tests on a glass melting furnace at Indiana
Glass Company in Dunkirk, Indiana.  These tests were conducted
from May 17 through 19, 1984, as part of a larger study designed
to evaluate the sampling and analytical procedures for measuring
inorganic arsenic from stationary sources.  Proposed Method 108*
provides total arsenic results  (particulate plus gaseous frac-
tion) .
 40 CFR 61, Appendix B, Method 108, July 1983.
                               1-1

-------
     The primary objective of this test program was to determine



the precision of proposed Method 108.  Relative standard devia-



tions (the standard deviation expressed as a percent of the mean



value) of four-train (quad) sample runs were used to estimate



method precision.  A total of nine quad train runs representing



36 individual samples were conducted using Method 108 sampling



and analytical procedures as described in the Quality Assurance



Project Plan developed and submitted in January 1984 to the EPA



Environmental Monitoring Systems Laboratory.  These data are



summarized in a report issued to EMSL-QAD.



     In this specific portion of the test program, four quad-



train tests were conducted using Method 108 procedures except



that probe and filter temperatures were elevated to approximately



204°C and 288°C in order to evaluate the effects of increased



sampling train temperature on arsenic distribution in the sam-



pling train.  During these runs, a single Method 108 sampling



train (121°C) was run for reference purposes.  Three Method 108



traverse tests were also conducted to provide additional data in



support of the arsenic standards developed to date.



     Section 2 summarizes and discusses the test results; Section



3 addresses quality assurance considerations specific to this



project; Section 4 describes the sampling locations and test



procedures; and Section 5 describes source operation.  Appendix A



presents sample calculations and computer printouts; Appendices B



and C contain the field data sheets and laboratory analytical
                               1-2

-------
results, respectively; Appendix D details the sampling and ana-



lytical procedures; Appendix E summarizes equipment calibration



procedures and results; and Appendix F contains a process descrip-



tion and the furnace operating data for the test period.
                                1-3

-------
                            SECTION 2



             SUMMARY AND DISCUSSION OF TEST RESULTS






2.1  SAMPLING AND ANALYTICAL PROTOCOL



     A four-train (quad) sampling system was used to collect



samples at the furnace exit stack.  This system allows four



trains to sample simultaneously at essentially a single point in



the stack (see Section 4).



     Because this sampling approach allows simultaneous sampling



at essentially a single point, it reduces the effect of varia-



tions in the velocity and particulate profiles on the sampling



results.  It also permits a statistically significant number of



samples to be taken in a short amount of time.  Further, since



two of the four trains are identical for every run, the within-



train precision can be determined at the same time as the rela-



tionship of the different trains is being compared.



     The Quad runs conducted were designed to evaluate the effect



of arsenic collection at elevated sampling temperatures.  Two of



the trains were heated to approximately 204°C (400°F) and two



trains were heated to approximately 288°C (550°F) for comparative



purposes.  Additionally, in three of the four quad tests con-



ducted, backup filters were maintained at approximately 121°C



(250°F) prior to the impinger section.
                                2-1

-------
     Figure 2-1 depicts the quad train configuration used for



these tests.  Individual train components were recovered and



analyzed for arsenic separately to evaluate the distribution of



arsenic in the sampling train.  In each train, the contents of



the first and second impingers were recovered, combined, and



analyzed for arsenic and the third and fourth impingers were



recovered, combined, and analyzed for arsenic.  The probe rinse



and front filter were recovered and analyzed according to pro-



cedures defined in Method 108.  In Trains A and D, the back-half



glassware of the front filter, the glass connector, and the



front-half glassware of the backup filter were rinsed with 0.1 N



NaOH and this rinse was analyzed for arsenic.  The backup filter



was analyzed separately in each case.



     During these runs, a single Method 108 sampling system



(designated RT) (121°C) was run for reference purposes.  Three



multipoint traverse tests utilizing a single Method 108 train



were also conducted at the completion of the quad-train tests.



     In each train, the probe and filter temperatures were set at



a predetermined temperature and monitored using multiterminal



digital indicators with thermocouple leads located in each probe



and immediately behind the Method 5 filter frits.






2.2  TEST RESULTS—ELEVATED TEMPERATURE RUNS



     Table 2-1 summarizes sampling conditions for the quad train,



reference train, and traverse train  (designated CD) test runs.



Table 2-2 summarizes the arsenic analytical results by sample
                               2-2

-------
BACKUP
METHOD 5
FILTER
                                   IMPINGER
                                    TRAINS
                        (204°C)                    (288°C)
                                   HEAT BOXES
              BACKUP
             'METHOD 5
              FILTER
              (12TC)
                                                    Az

                                                    B-
                                                       FRONT VIEW
                                     B A D C
                                     CO CO CD CO
                                     oo oo
                                     er oc a; of.
                                     D. Q- CL O.
OD  OA

oC  OB
                                                        BACK VIEW
         Figure  2-1.   Quad  train  system  for  elevated  temperature  tests,
                                     2-3

-------
TABLE 2-1.  SUMMARY OF SAMPLE CONDITIONS
Run
No.
IDA
10B
IOC
10D
10-RT

11A
11B
11C
11D
11 -PT

1ZA
12B
12C
120
12-RT

13A
13B
13C
13D
13-RT

CD-I

CO-2

CD-3

Sampling
type
Modified
Method 108


Method 108

Modified
Method 108


Method 108

Modified
Method 108


Method 108

Modified
Method 108


Method 108

Method 108

Method 108

Method 108

Date (1984)
and
time (24-h)
5/17
11:45-12:55


5/17
11:45-12:55
5/17
17:11-18:21


5/17
17:11-18:21
5/18
10:59-12:09


5/18
10:59-12:09
5/18
15:36-16:46


5/18
15:36-16:46
5/19
10:32-11:45
5/19
13:05-14:18
5/19
15:12-16:22
Metered
volume,
dsm3
1.07
1.12
1.18
1.09
1.41

1.02
1.12
1.14
1.00
1.33

1.23
1.33
1.34
0.29
1.53

1.14
1.28
1.30
1.13
1.37

1.03

1.06

1.07

Isoki-
netic, %
101.0
97.5
101.2
100.4
96.0

101.2
97.8
100.2
99.3
98.7

107.1
103.0
101.2
101.2
99.8

94.5
98.2
101.4
100.0
100.6

105.6

102.1

100.3

Mois-
ture, *
8.4
8.9
8.4
8.6
8.1

8.5
6.9
8.3
8.8
8.4

8.6
8.4
8.6
9.1
8.5

8.9
8.7
8.6
9.7
9.0

8.0

8.3

7.7

Sampling conditions
Gas
temper-
ature, °C
298
298
298
298
301

301
301
301
301
311

286
286
286
286
292

274
274
274
274
301

260

255

248

Probe
temper-
ature, °C
208
206
248
178
121

193
200
295
286
121

205
210
279
174
121

208
209
281
279
121

147

188

180

Filter
temper-
ature, °C
204
206
284
286
121

201
210
262
285
124

208
210
261
284
122

206
209
239
263
122

143

124

123

Backup fil-
ter temper-
ature, "C
91
NA
NA
120
NA

NA
NA
NA
NA
NA

118
NA
NA
104
NA

122
NA
NA
121
NA

NA

NA

NA

                  2-4

-------
                                                  TABLE 2-2.    SUMMARY  OF  ARSENIC  ANALYTICAL  RESULTS
                                                                QUAD  AND REFERENCE  TRAIN RUNS
ro
 i
en
Run
No.
10A
10B
IOC
10D
10-RT
11A
11B
11C
11DC
11-RT
12A
12B
12C
12Dd
12-RT
13A
13B
13C
13D
13-RT
Sample
volume,
dNm3
1.07
1.12
1.18
1.09
1.41
1.02
1.12
1.14
1.00
1.33
1.23
1.33
1.34
0.292
1.53
1.14
1.28
1.30
1.13
1.37
Arsenic sample weights, mg
Probe
rinse
0.323
0.686
0.483
0.775
1.09
0.792
0.666
0.618
0.468
1.21
0.280
0.786
0.490
0.627
1.11
0.926
0.866
0.845
0.786
1.31
Front3
filter
8.50
9.26
9.81
9.81
11.17
7.57
9.19
8.98
1.75
11.17
10.48
10.86
10.96
2.05
10.79
9.26
10.55
10.35
9.55
11.50
Total
front
half
8.82
9.95
10.29
10.59
12.26
8.36
9.86
9.60
2.22
12.38
10.76
11.65
11.45
2.68
11.90
10.19
11.42
11.20
10.34
12.81
Back-half5
Glass
connector
0.05
NA
NA
0.05
NA
NA
NA
NA
NA
NA
0.05
NA
NA
0.05
NA
0.05
NA
NA
0.05
NA
Backup
filter
0.05
NA
NA
0.05
NA
NA
NA
NA
NA
NA
0.05
NA
NA
0.05
NA
0.05
NA
NA
0.05
NA
Impingers
1 & 2
0.05
0.05
0.05
0.05
0.05
0.05
0.05
0.05
2.38
0.05
0.05
0.05
0.05
0.05
0.05
0.05
0.05
0.05
0.05
0.10
Impingers
3 & 4
0.05
0.05
0.05
0.05
NA
0.05
0.05
0.05
0.21
0.05
0.05
0.05
0.05
0.05
0.05
0.05
0.05
0.05
0.05
NA
Total
back half
0.05
0.05
0.05
0.05
0.05
0.05
0.05
0.05
2.59
0.05
0.05
0.05
0.05
0.05
0.05
0.05
0.05
0.05
0.05
0.10
Concentration,
ng/dsm3
Front
half
8.24
8.88
8.72
9.72
8.70
8.20
8.80
8.42
2.72
9.31
8.75
8.76
8.54
9.18
7.78
8.94
8.92
8.62
9.15
9.35
Back
half
-
-
-
-
-
-
-
-
2.59
-
-
-
-
-
-
-
-
-
-
0.07
Total
train
8.24
8.88
8.72
9.72
8.70
8.20
8.80
8.42
4.81
9.31
8.75
8.76
8.54
9.18
7.78
8.94
8.92
8.62
9.15
9.42
                     The front filter data include the Parr bomb results, which constituted approxmately 1 percent of the total arsenic on
                     the filter.

                     NaOH rinse and impinger solution blank values  ranged from 0.0 to 0.05 mg,  therefore, back half values less than 0.05
                     mg are not reported.

                    cRun 11D is considered void due to a  ruptured filter frit and subsequent loss of sample (see Pages 2-8 and 2-12).

                     Run 120 was terminated approximately 14 minutes into the test due to a ruptured filter frit support (see Page 2-12).

-------
fraction, and Table 2-3 presents statistical data for the grouped




quad runs.



     Sample volumes were consistent and ranged between 1.00 and



1.34 dsm3 for the quad runs conducted during the full test



period.  Quad Run 12D was terminated approximately 14 minutes



into the test due to a broken filter frit; the sample volume for



this run was 0.292 dsm3.  Sample volumes for the reference train



tests ranged between 1.33 and 1.53 dsm3.  Isokinetic sampling



rates ranged from 96.0 to 107.1 percent, which is within the



acceptable range of 90 to 100 percent.  The probe and filter



temperatures represent average values determined from data



recorded on the field data sheets.  The desired temperature for



paired Trains A and B was 204°C and for paired Trains C and D,



288°C.  As shown, filter temperatures for Trains A and B ranged



from 201°C to 210°C and the probe temperatures ranged between



193° and 210°C.  In Trains C and D, the filter temperatures were



more variable ranging between 239° and 286°C, and the probe



temperatures ranging from 174° to 295°C.  The backup filter



temperatures in Runs 10A and 10D, 12A and 12D, and 13A and 13D



ranged from 91° to 122°C.  No backup filters were utilized for



Quad Run 11.  In each quad test, the reference train probe and



filter temperature was maintained at approximately 121°C.



     The moisture content of the stack gas was generally con-



sistent  in each run, and the average gas temperatures ranged from



274° to  310°C.



     As  shown in Table 2-2, arsenic sample weights are reported



in milligrams  (mg) for each sample fraction analyzed.  The front




                               2-6

-------
          TABLE 2-3.  STATISTICAL DATA FOR GROUPED RUNS
Quad Run
No.a
10A
10B
IOC
10D
11A
11B
11C
11D
12A
12B
12C
12D
13A
13B
13C
13D
Overall
means
Individual
run value
8.24
8.88
8.72
9.72
8.2
8.80
8.42
4.81
8.75
8.76
8.54
9.18
8.94
8.92
8.62
9.15


Group, mean
fcb

8.89


8.47


8.68


8.91

8.74e

°' ,
mg/dsm

0.617


0.304


0.124


0.218

0.367f

RSD,d
%

6.9


3.6


1.4


2.4

4.209

 Sample Nos. 11D and 12D are considered invalid and are not included
 in the group data.

 Mean concentration.

cWithin-run standard deviation with N-l weighting for sample data.

 Within-run relative standard deviation is the standard deviation
 expressed as a percent of the mean concentration.
p
 Simple averages of tabulated data.
 Pooled standard deviation;\J  £o .

              ~
                               2-7

-------
filter weight includes results for both the NaOH extract and the



Parr bomb (HF/HNO-) extract.  The extract results constituted



approximately 1 percent of the total arsenic on the filter.



     Arsenic was found mainly in the front half (probe and fil-



ter) of each of the 16 individual trains with the exception of



Run 11D.  During this run, the filter frit support ruptured but



sampling continued until completion of the run.  Obviously, some



arsenic was carried to the back half of the sampling train as



evidenced by the reported weight (2.59 mg) in the back half.  Run



11D is considered an invalid sample and is not included in any of



the grouped averages or statistical calculations.



     The total amount of arsenic found in the front half was 99



percent in each case, and at least 90 percent of this amount was



found in the filter fraction.  No significant amount of arsenic



was present in any of the back-half components.  The 0.05 mg



limit reported in Table 2-2 was established after careful anal-



ysis of the sample "blank" data.  These data are summarized in



Section 3 of this report.  In summary, 60 percent of the blank



values for the NaOH rinse and H-O impinger solutions were at or



below the analytical detection limit  (0.002 to 0.005 mg); the



remaining blank values ranged up to a maximum value of 0.05 mg.



Values below 0.05 mg were considered insignificant because the



back-half arsenic content constituted less than 0.5 percent of



the total arsenic collected, the liquid fraction blank data were



variable, and 8.4 mg was the minimum amount of arsenic collected



in any one train run for at least 60 minutes.  Note that in Run
                              2-8

-------
12D, the total arsenic collected was 2.7 mg.  The filter frit



support ruptured approximately 14 minutes into the test and



sampling was immediately terminated.  The train was disassembled



and recovered according to routine procedures.  No significant



amount of arsenic was present in the back half of the train in



this run, and since the concentration is comparable with the



within-run data, the sample run is considered representative.



Because only 0.292 dsm3 was metered, however, the sample volume



does not conform to Method 108 specifications; therefore, the



concentration value is not included in the group statistical data



presented in Table 2-4.



     The statistical data presented in Table 2-3 are comparable



with data obtained during the EMSL-QAD portion of this test pro-



ject.  Statistical data for nine Method 108 quad train runs  (36



individual samples; 121°C sample temperature) showed an overall



mean of 9.59 mg/dsm3 with mean arsenic concentrations of indi-



vidual quad runs ranging from 8.48 to 10.55 mg/dsm3.  The stan-



dard deviations of the EMSL tests ranged from 0.10 to 1.45



mg/dsm3 with a pooled mean value of 0.59 mg/dsm3.  The mean



relative standard deviation  (RSD) for the nine runs was 6.14



percent.  Table 2-4 summarizes the EMSL quad train results.



     The standard deviations of the elevated temperature quad



runs ranged from 0.124 to 0.617 mg/dsm3 with a pooled mean value



of 0.367 mg/dsm3.  The RSD values ranged from 1.4 to 6.9 percent



with a mean RSD of 4.2 percent.  The mean arsenic concentration



of the individual quad runs ranged from 8.20 to 9.72 mg/dsm3 with
                                2-9

-------
  TABLE 2-4.   STATISTICAL  DATA  FOR GROUPED RUNS - EMSL QUAD TRAIN TESTS
Quad
run
No.
1A
IB
1C
ID
2A
2B
2C
2D
3A
3B
3C
3D
4A
4B
4C
4D
5A
SB
5C
5D
6A
6B
6C
6D
7A
7B
7C
7D
Concentration, mg/dsm3
Individual
run value
9.11
8.87
6.35
9.58
8.82
10.05
9.22
9.28
9.26
10.24
10.13
10.04
9.53
9.28
8.73
9.17
10.62
10.53
10.62
10.42
9.93
9.96
9.92
10.48
9.57
9.13
9.95
9.82
Groupamean

8.48


9.34


9.92


9.18


10.55


10.07


9.62

°-b,
mg/dsm3

1.45


0.514


0.446


0.334


0.10


0.272


0.361

RSD,C
%

17.1


5.5


4.5


3.6


0.95


2.7


3.8

(continued)
                                  2-10

-------
 TABLE  2-4  (continued)
Quad
run
No.
8A
8B
8C
80
9A
9B
9C
9D
Overal means
Concentration, mg/dsm3
Individual
run value
10.24
9.95
9.44
9.57
9.06
9.21
9.24
9.76
-
Group.mean

9.80


9.32

9.59d

a,
mg/dsm3

0.365


0.305

0.5896

RSD,C
%

3.7


3.3

6.14f
 Mean concentration.
 Within-run standard deviation with N-l  weighting  for sample  data.
cWithin-run relative standard deviation is the standard  deviation expressed
 as a percent of the mean concentration.
 Simple averages of tabulated data.
ePooled standard deviation;
 RSD
         10 -/X.
                              n
n
                                     2-11

-------
an overall mean of 8.74 mg/dsm3.  The overall mean of the ele-



vated temperature runs compares to within 10 percent of the



overall mean of the EMSL quad runs.



     As shown in Table 2-2, the Method 108 reference train tests



that ran concurrently with the quad train tests are comparable



relative to total arsenic concentration and distribution.  Over-



all, there is less than a 10 percent difference between the quad



run means and the reference train arsenic concentrations.  In Run



10, the quad group mean was 8.89 mg/dsm3 compared with the refer-



ence train value of 8.70 mg/dsm3.  In Run 11, the quad group mean



excluding 11D was 8.47 mg/dsm3 and the reference train value was



9.31 mg/dsm3.  The group mean in Run 12 excluding 12D was 8.68



mg/dsm3 and the reference train value was 7.78 mg/dsm3.  In Run



13, the quad group mean was 8.91 mg/dsm3 compared with 9.42



mg/dsm3 for the reference train.



     These data in conjunction with the EMSL quad run data sug-



gest no significant difference between arsenic concentrations



measured by Method 108 (121°C) and modified Method 108 (elevated



sample temperature) at this source.  The data also indicate no



significant difference between samples collected at 204°C and



those collected at 288°C.



     Several factors observed during this test series that could



have affected sample results are addressed as follows.  The



filter frit supports for Quad Runs 11D and 12C and 12D ruptured



during testing.  The ruptures were attributed to the deteriora-



tion of the silica rubber gasket due to the sample temperature
                               2-12

-------
(288°C) and subsequent failure under vacuum.  As mentioned in



Subsection 2.2, Run 11D was not terminated when the gasket



failed; consequently, a significant amount of arsenic was found



in the back half of the sampling train.  The arsenic measured by



Train 11D, however, did not compare with the within-run samples



on a total weight basis.  Because particulate was noticeable on



the frit, it was rinsed with 0.1 N NaOH into the container hold-



ing the contents of the first two impingers.  Sample fractions



from this run were reanalyzed and rechecked; no discrepancies



were found in the reported analytical data.



     Runs 12C and 12D were terminated immediately upon rupture of



the frit.  In Train 12C, the rupture occurred with less than one



minute to go in the 70-minute test; in Train 12D, the rupture



occurred approximately 14 minutes after the start of the test.



In each case, no significant amount of arsenic was found in the



back half of the train.  Also, the arsenic concentrations deter-



mined in these two runs are comparable with the within-run data.



New filter frits were used in each quad run, thereby minimizing



this problem.



     Several back-half samples  (connector glassware and backup



filters) from Quad Runs 10, 11, and 12 were contaminated by a



brown, oily substance believed to be volatilized probe heat tape



glue.  This phenomenon is attributed to the high temperatures to



which the probes were heated and the use of an asbestos string



gasket material at the nozzle end of the probe.  When heated to



288°C, the heat tape used in the construction of the probe heat-



ing system burned resulting in the volatilization of the tape




                                2-13

-------
glue.  A visual inspection of the affected probes showed a heavy



deposit of contaminant on the asbestos gaskets as well as



distinct trails of the contaminant on the nozzle end of the glass



liner.



     A heavier disposition of the contaminate was observed on the



trains heated to 288°C than those heated to 204°C.  The material



was recovered by using the 0.1 N NaOH rinse and a nylon brush for



each affected sample fraction.  Since the material was recovered



and digested according to Method 108 in a Parr bomb, any arsenic



from the gas stream that might become bound to the material would



be analyzed thus precluding a low bias on sample results.  As the



back-half results indicate (Table 2-2), no arsenic was found in



these "contaminated" sample fraction; thus any bias in arsenic



measurements is believed to be minimal.



     Another phenomenon associated with the filter frit support



occurred during the 288°C runs.  Experiments conducted in our



laboratory showed that the standard glass frit filter support



with a silicon rubber gasket could withstand temperatures up to



260°C.



     Above 260°C, deterioration of the gasket was noticeable as



evidenced by a light film of material on the filter holder glass-



ware.  The material was believed to be a form of silicon oxide.



This same white material was present on all of the 288°C filter



glassware.  A recovered sample from the laboratory experiment was



analyzed for arsenic.  A detection limit of 0.003 milligrams per



liter was established for this sample and no detectable arsenic



was found.  The data indicate that this material would not cause




                               2-14

-------
a high bias in arsenic results from these runs.  Since the mate-



rial is recoverable and would be digested in a Parr bomb, any



arsenic from the gas stream which might become bound to the



material would be analyzed thus precluding a low bias on sample



results.






2.3  METHOD 108 TRAVERSE TEST RESULTS



     Table 2-5 summarizes the sample and flue gas conditions and



Table 2-6 presents the arsenic emissions data for the Method 108



traverse tests.



     Triplicate tests were conducted at the completion of the



quad train runs following procedures described in Method 108.



Twenty-four traverse points  (12 per port) were used to traverse



the cross-sectional area of the stack.  Each point was sampled



for 2.5 minutes yielding a total test time of 60 minutes.



     Sample volumes for the three tests were consistent and



ranged from 1.03 to 1.07 dsm3.  Isokinetic samples rates ranged



from 100.3 to 105.6 percent.  The flue gas volumetric flow in



actual cubic meters per minute  (m3/min) averaged 886 m3/min  (440



dsm3/min at 20°C and 760 mmHg).  The gas temperature and moisture



content averaged 254°C and 8.0 percent, respectively.  Flue gas



composition was determined by analyzing integrated bag samples



collected during each test with an Orsat gas analyzer.  Oxygen



 (Op), carbon dioxide (CO-), and carbon monoxide  (CO) contents



averaged 13.9, 3.8, and 0.0 percent, respectively.



     As presented in Table 2-6, the total arsenic catch in milli-



grams ranged from 8.67 mg for Test CD-3 to 9.65 mg for Test CD-2.





                               2-15

-------
                             TABLE 2-5.  SUMMARY OF SAMPLE AND FLUE GAS CONDITIONS
                                            ARSENIC TRAVERSE TESTS
Run No.
CD-I
CD-2
CD-3
Date
(1984)
5/19
5/19
5/19
Average
Mete red
volume,
dsm3
1.03
1.06
1.07
1.05
Moisture
content,
%
8.0
8.3
7.7
8.0
Stack gas
tempera-
ture, °C
260
255
248
254
Gas composition,3 %
02
14.8
12.1
14.8
13.9
C02
3.2
4.7
3.5
3.8
CO
0.0
0.0
0.0
0.0
Volumetric.
flow rate
m3/min
851
903
903
886
dsm3/min
418
446
455
440
Isoki-
netic, %
105.6
102.1
100.3
-
 Gas  composition  determined using an Orsat gas analyzer.
'Volumetric  flow  rate  in actual cubic meters per minute (m3/min) and dry standard  cubic  meters
 per minute  (dsmVmin).

-------
TABLE 2-6.  SUMMARY OF ARSENIC  ANALYTICAL RESULTS
                 TRAVERSE TRAIN
Run No.
CD-I
CD-2
CD-3
Metered
volume,
dsm3
1.03
1.06
1.07
Average
Sample
weight, mg
Total arsenic
9.60
9.65
8.67
9.31
Concentra-
tion, mg/dsm3
9.32
9.10
8.10
8.84
Mass emission
rate, kg/h
0.23
0.24
0.22
0.23
                        2-17

-------
Total arsenic concentration averaged 8.84 mg/dsm3 with a corre-



sponding average mass emission rate of 0.23 kilograms per hour



(kg/h).   These average results obtained by multipoint, isokinetic



traverse techniques are comparable to results obtained during the



quad train runs.






2.4  PROCESS SAMPLES



     Several finished glass samples were obtained during the test



program to determine the arsenic content on a weight basis.



Table 2-7 summarizes the process sample results.  Results were



consistent and the arsenic content by weight was approximately



0.05 percent.  The samples analyzed were drinking mugs and three



portions of each mug  (top, handle, and bottom) were analyzed.



Initially, glass chunks from the three sample fractions were



placed in Teflon bombs and digested using the Parr bomb procedure



from EPA Method 108.  After extended heating, the glass chunks



did not dissolve.  Additional glass fragments were ground in an



agate mortar and pestle and the resulting powder was placed in



Teflon bombs and digested per the Method 108 Parr bomb procedure.



After extended heating, a white precipitate remained in the bomb.



The sample was filtered through a Teflon filter and the filtrate



was analyzed for arsenic per Method 108.  The remaining precipi-



tate was gelatinous in nature rather than a dense powder of the



original sample.  The precipitate was redigested using a Parr



bomb and the resulting solution was analyzed for arsenic per
                               2-18

-------
TABLE 2-7.   PROCESS  SAMPLE ANALYTICAL RESULTS
Sample type
Drinking mug
Drinking mug
Lab No.
DM924
DM924
DM924
DM929
DM929
DM929
DM929R
Description
5/14; #2535 (handle)
5/14; #2535 (top)
5/14; #2535 (bottom)
5/19; #2540 (handle)
5/19; #2540 (top)
5/19; #2540 (bottom)
5/19; $2540 (bottom)
Total arsenic,
% by weight
0.052
0.055
0.059
0.055
0.059
0.055
0.058
                     2-19

-------
Method 108.  This fraction consistently contained approximately 1



percent of the amount of arsenic found in the original filtrate.



These numbers were combined and reported.
                               2-20

-------
                            SECTION 3



                    PROJECT QUALITY ASSURANCE






     Because the desired end product of testing is to achieve



representative emission results, quality assurance is one of the



main facets of stack sampling.  Quality assurance guidelines



provide the detailed procedures and actions necessary for defin-



ing and producing acceptable data.  Five such documents were used



in this test program to ensure the collection of acceptable data



and to provide a definition of unacceptable data.  The following



documents comprise the Quality Assurance Project Plan prepared by



PEI and reviewed and approved by the Environmental Monitoring



Support Laboratory of the EPA  (see Volume II - Appendix F); the



detailed site test plan prepared by PEI and reviewed by the



Emission Measurement Branch; the EPA Quality Assurance Handbook



Volume III, EPA-600/4-77-027; the PEI Emission Test Quality



Assurance Plan; and the PEI Laboratory Quality Assurance Plan.



The last two, which are PEI's general guideline manuals, define



the company's standard operating procedures and are followed by



the emission testing and laboratory groups.



     In this specific test program, the following steps were



taken to ensure that the testing and analytical procedures



produced quality data.
                               3-1

-------
     0    Calibration of all field sampling equipment.

     0    Checks on train configuration and calculations.

     0    Onsite quality assurance checks (i.e., leak checks of
          the sampling train, pitot tube, and Orsat line) and
          quality assurance checks of all test equipment prior to
          use.

     0    Use of designated analytical equipment and sampling
          reagents.

     0    Internal and external audits to ensure accuracy in
          sampling and analysis.

     Table 3-1 lists the sampling equipment used to perform the

arsenic tests and the calibration guidelines and limits.  In

addition to the pre- and post-test calibrations, a field audit

was performed on the metering systems and thermocouple digital

indicators used in the sample runs.  PEI-constructed critical

orifices were used in the dry gas meter audits.  The onsite

audits were made at the beginning, middle, and end of the test

program.   Figures 3-1 through 3-23 present the results of the

pre-test, mid-test, and post-test onsite audits.  These data were

used to assess the operational status of the sampling equipment

relative to guidelines established by the U.S. EPA.  The results

of the three field audits indicate that the sample equipment was

functioning properly throughout this test series.

     Figure 3-24 is an example of an unacceptable meter box

audit.  The audit value for the meter coefficient deviation was

greater than ±5 percent, which was considered unacceptable by the

PEI Project Manager; therefore, the meter box was not used for

this test program.
                                3-2

-------
                                         TABLE 3-1.   FIELD  EQUIPMENT CALIBRATION
Equipment
Meter box













Pitot tube

Digital
indicator
Thermocouple


ID No.
FB-1 Train A


FB-5 Train B


FB-8 Train D


FB-10 Train C


FB-9 (Traverse
tests)
FB-9 (Reference
train tests)
513
514
508
220
221
411 - (stack)
412 - (stack)
601 - Probe
605 - Filter
Calibrated
against
Wet test meter













Standard pitot
tube

Millivolt
signals
ASTM-3F


Allowable error
Y ±0.02 Y
AH (3 ±0.15
(Y ±0.05 Y post-test)











Cp ±0.01

0.5%
1.5%
(±2% saturated)


Actual
error
0.004
0.09
+1.35
0.002
0.05
+0.10
0.006
0.07
-0.20
0.012
0.07
-0.52
0.013
0.08
+1.30
-
-
0.2%
0.4%
0.4%
0.2%
0.8%
1.0%
Within
allowable
limits
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
OK
OK

X
X
X
X
X
X
Comments














Visually
inspected
on site




OJ
I
U)
    (continued)

-------
TABLE 3-1 (continued)
Equipment
Thermocouple
(cont'd)




Orsat analyzer

Impinger
thermometer



Mettler
balance
Barometer
Dry gas
thermometer




ID No.
614 - Probe
615 - Probe
616 - Probe
618 - Probe
619 - Probe
620 - Filter
141

433
434
435
446
385
M-l
407
FB-1
FB-5

FB-8

Calibrated
against





Standard gas

ASTM-3F



Type S weights
NBS traceable
barometer
ASTM-3F




Allowable error





±0.5%

±2°F



±0.5 g
+0.10 in.Hg.
(0.20 post-test)
±5°F




Actual
error
0.4%
0.6%
0.5%
0.6%
0.6%
1.2%
0.0%
0.2%
0.0%
1.2°F
0.5°F
1.0°F
1.0°F
1.0°F
+0.1 g
0.01
in.Hg.
4°F
5°F
4°F
2°F
2°F
3°F
Within
allowable
limits
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
Comments





co2
CO






Inlet
Outlet
Inlet
Outlet
Inlet
Outlet
(continued)

-------
TABLE 3-1 (continued)
Equipment
Dry gas
thermometer
(cont'd)


Probe nozzle












ID No.
FB-10


FB-9

1A
IB
1C
ID
2A
2B
2C
2D
5-106 (RT
tests)
2-117
(Traverse
tests)
Calibrated
against





Caliper












Allowable error





Dn ±0.004 in.












Actual
error
2°F
2°F

2°F
2°F
0.001 in.
0.002 in.
0.000 in.
0.001 in.
0.002 in.
0.002 in.
0.000 in.
0.002 in.
0.001 in.

0.003 in.


Within
allowable
1 imits
X
X

X
X
: X
X
X
X
X
X
X
X
X

X


Comments
Inlet
Outlet

Inlet
Outlet














-------
                       FIELD AUDIT REPORT:  DRY GAS METER
                               BY CRITICAL ORIFICE
DATE:
         -5-
                                        CLIENT;
BAROMETRIC PRESSURE (P):
-------
                       FIELD  AUDIT  REPORT:  DRY GAS METER
                               BY CRITICAL ORIFICE
DATE:
                                        CLIENT:
BAROMETRIC PRESSURE (Pbar);^go 1n.Hg  METER BOX NO.
ORIFICE NO.
ORIFICE K FACTOR:
                            X/£>
                                         PRETEST

                                         AUDITOR:
                                                                         1n.H0
Orifice
manometer
reading
AH,
1n.H20
P05
Dry gas
meter
reading
ft3
^tf.fcc
;2.6>3,60c
Temperatures
Ambient
Tai/Taf
°F
•^o
?o
Average
V
°F
10
Dr
Inlet
°F
•^
&o
/gas meter
Outlet
°F
1^
^
-------
                       FIELD AUDIT REPORT:  DRY GAS METER
                              BY CRITICAL ORIFICE
DATE:
                                        CLIENT:
BAROMETRIC PRESSURE (Pbar):Pi££ 1n.Hg  METER BOX NO

ORIFICE NO.   3	       PRETEST
                            ''    "^      AUDITOR*
ORIFICE K FACTOR:
                                                                   /f/  1n.H20
Orifice
manometer
reading
  AH,

1n.H,0
           Dry gas
           meter
           reading
           w
             ft3
                                     Temperatures
                          Ambient
                     Ta1/Taf


                       °F
Average
                 Dr
 Inlet

T11/Tif

  °F
         / gas meter
Outlet
Average
Duration
   of
   run
    0
   m1n.
                                                             7/.S
Dry gas
meter
V ft3
/3.200
V|nstd'
ft3
/P.?*/
Vmact»
ft3
/P 5? ?
Audit,
Y
O3*&-
Y
devia-
tion, %
dP^I
Audit
AH(a,
1n.H20
/7V
AH@ Devia-
tion, 1n.H20
<9.07
 m
  std
            17.647(Vm)(Pbar + AH/13.6)
 m
   act
            1203( 0 )( K )(Pbflr)
               (Ta + 460)
Audit Y "
             m
              std
                                                                   x 100
 Audit  AH(? =  (0.0317)(AH)(Pbar)(Tm + 460)
                                                    0
                                                                           1n.H0
 Audit  Y must be  in the range, pre-test Y ±0.05 Y.
 Audit  AHP must be in the range pre-test AH@ ±0.15 inches
                  Figure 3-3.   Pre-test audit report:  dry gas meter  by
                        critical orifice  (Meter Box FB-8, Tram D).
                                     3-8

-------
                      FIELD AUDIT REPORT:  DRY GAS METER
                              BY CRITICAL ORIFICE
DATE:
                                        CLIENT:
BAROMETRIC PRESSURE (Pbar): "Z?'*° in.Hg  METER BOX NO.
ORIFICE NO.      /2- _  PRETEST
ORIFICE K FACTOR:   ^."R^S^O'^     AUDITOR;/Su'
                                                              SO  ~
                                                              AHf  /. #/  ,1n.H,0
Orifice
manometer
reading
AH,
1n.H20
/ft?
Dry gas
meter
reading
VV
ft3
O^o. too
031. <1*o
Temperatures
Ambient
Ta1/Taf
°F
•}o
?l
Average
T,-
•F
10.6
Dr
Inlet
°F
^(s>
^
/gas meter
Outlet
VV'
7z-
^
Average
°F
H&
Duration
of
run
0
min.
*2 I •*"*"
Dry gas
meter
V ft3
/1,00V
v
std'
ft3
/6>'?>q3
v
act*
ft3
K,414

Audit,
Y
oj1*

AH@ Devia-
tion, 1n.H20
0.0(0
 m
  std
            17.647(Vm)(Pbar * AH/13.6)
                   (Tm + 460)
 m
  act
            1203( 0 )( K )(Pbflr)
               (Ta + 460)
                          1/2
Audit Y
             m
              'act
             m
              'std
                              Y deviation
Audit Y - Pre-test Y
        Audit Y
x 100 «= -
Audit AH(? = (0.0317)(AH)(Pbar)(Tm + 460)
                                                        AH/13.6)
Audit Y must be 1n the range, pre-test Y ±0.05 Y.
Audit AH@ must be In the range pre-test AH@ ±0.15 Inches H20.


              Figure 3-4.  Pre-test audit report:   dry gas  meter  by
                    critical  orifice (Meter Box FB-10, Train  C).
                                     3-9

-------
                       FIELD AUDIT REPORT:   DRY  GAS METER
                               BY CRITICAL  ORIFICE
DATE:
                                         CLIENT:
BAROMETRIC PRESSURE (PhaJ: P^£2_in.Hg   METER  BOX  NO.
ORIFICE NO.
                      bar
ORIFICE K FACTOR:
                                         PRETEST  Y:.

                                         AUDITO
Orifice
manometer
reading
AH,
in.H20
->,zs
Dry gas
meter
reading
vv
ft3
OT*.Ooo
0^/3,3-00
Temperatures
Ambient
Tai/Taf»
°F
Tz_
^3^
Average
V
&•
Dry gas meter
Inlet
VTif
°F
^Z-
^ ^
Outlet
w-
°F
^Z-
^
Average
°F
^.5
Duration
of
run
0
min.
ZL'S <&&
Dry gas
meter
v ft3
70.-2.
v
std'
ft3
PO./«
v
act*
ft3
19.117,

Audit,
Y
0*50
Y
devia-
tion, %
-3.15
Audit
AH@,
in.H20
,2.06

AH@ Devia-
tion, in.H20
0.02, .
 m
  std
            17.647(Vm)(Pbar + AH/13.6)
 m
  act
            1203( 0 )( K )(Pbar)
               (Ta + 460)
                         1/2
Audit Y =
             m
              'act
             m
Y deviation =
Audit Y - Pre-test Y
        Audit Y
                                                                    x  100  =  -3.
              std
Audit AH@ = (0.0317)(AH)(Pbar)(Tm + 460)
                                          Y TVJ(Pbar + AH/13.6)
                                             in.H20
Audit Y must be in the range, pre-test Y ±0.05 Y.
Audit AH@ must be in the range pre-test AH@ ±0.15 inches H^
          Figure 3-5.  Pre-test audit report:  dry gas meter by critical
            orifice (Meter Box FB-9, Single Point-Traverse Tests).
                                     3-10

-------
                THERMOCOUPLE DIGITAL INDICATOR
                        AUDIT DATA SHEET
Date 6~~ //'
Indicator No.
                                                 Operator
Test Point
No.
1
2
3
4
Millivolt
signal*




Equivalent
temperature,
°F*
32-
2-OO
5^0
/1 44
Digital Indicator
temperature reading,
°F
3,2-.
203~
*S^o
/Mb
Is
Difference,
I
G*O
- o.s<^
o.o
-o.(Z_
Percent difference must be less than or equal  to  0.5%.

Percent difference:

  (Equivalent  temperature °R* Digital Indicator  temperature reading °R)(100%)
                             (Equivalent temperature  R)
Where °R « °F * 460°F
  These values  are to be obtained from the calibration data sheet for the
  calibration device.
              Figure 3-6.  Pre-test thermocouple digital indicator
                     audit data sheet (Indicator No. 220).
                                     3-11

-------
                THERMOCOUPLE DIGITAL INDICATOR
                        AUDIT DATA SHEET
Date ^ - //• *?T   Indicator No.    P«2-/      Operator
                                                            zz
Test Point
No.
1
2
3
4
Millivolt
signal*




Equivalent
temperature,
°F*
33-
3-&O
-^o
//fy
Digital Indicator
temperature reading,
•F
So
c2-0 1
54o
/Ifa
Difference,
X
0,41
-0./5
O.o o
— 0. 1^
Percent difference must be less than or equal  to 0.52.
Percent difference:
  (Equivalent temperature °R- Digital Indicator temperature reading °R)(100'%)
                             (Equivalent temperature^R)
Where °R • °F + 460°F
  These values are  to be obtained from the calibration data sheet for the
  calibration device.
             Figure 3-7.   Pre-test thermocouple digital indicator
                    audit  data sheet (Indicator No. 221).
                                     3-12

-------
Audit Name
                          ON-SITE AUDIT DATA SHEET
Date:
Auditor:
"— 	 ' 	 	 	 	 	 7^ 	
Equipment
*%?, Meter box
£?' Inlet thermo.
poO „
F6Z- Meter box
& ' outlet thermo.
£AS
/ *-*^
<*
?Z
^/
62
^
1>V
(01:
&a
£&
6?
(ol-

#A

Deviation
3
Z
*
2
*/
2~
Ji
1
z
z
?
3

X4

Max. fcflowable
Deviation
5°F
5°F
2°F
7°F
See table
0.7%
0.5 grams
0.20 in. Hg
Reference temp. °F
Max. deviation °F
32-140
7
141-273
9
274-406
11
407-540
13
541-673
15
674-760
17
 * Correction factor:

     NWS value (1n. Hg) - [Altitude (ft)/1000(ft/in. Hg)] + 0.74 1n. Hg**

** 0.74 1n. Hg Is the nominal correction factor for the reference barometer
   against which the field barometer was calibrated.

If 1t is not feasible to perform the audit on any piece of equipment, record
"N/A" in the space provided for the data.



              Figure  3-8.  Pre-test onsite audit data sheet.
                                     3-13

-------
                          ON-SITE AUDIT DATA SHEET
Audit
Equipment
^'?Meter box
F&1> Inlet thermo.
£'° Meter box
pd t outlet thermo.
r*35 Impinger
ua< thermometer
i-?-*
Stack
thermometer
or
Thermocouple
Orsat
analyzer
Trip
balance
Barometer
Reference
ASTM-3F at
ambient temp.
ASTM-3F at
ambient temp.
ASTM-3F at
ambient temp.
ASTM-3F at
ambient temp.
ASTM-3F at
stack temp.
% 02 in
ambient air
IOLM std.
weight
Corrected*
NWS value
Reference
Value
%
Gf ^m
+ ^7
t<*


20.8%


Value
Determined
^
<*><<>
r?





Deviation
3
2-
O





Max. Allowable
Deviation
5°F
5°F
2°F
7°F
See table
0.7%
0.5 grams
0.20 in. Hg
Reference temp. °F
Max. deviation °F
32-140
7
141-273
9
274-406
11
407-540
13
541-673
15
674-760
17
 * Correction factor:

     NWS value (in. Hg) - [Altitude (ft)/1000(ft/in. Hg)] + 0.74 in. Hg**

** 0.74 in. Hg is the nominal correction factor for the reference barometer
   against which the field barometer was calibrated.

If 1t is not feasible to perform the audit on any piece of equipment, record
"N/A" in the space provided for the data.
              Figure 3-9.  Pre-test onsite audit data  sheet.
                                    3-14

-------
                       FIELD AUDIT REPORT:  DRY GAS METER
                               BY  CRITICAL ORIFICE
DATE:
                                         CLIENT:
BAROMETRIC PRESSURE (Pbar):P?y^2-1n.Hg  METER  BOX NO

ORIFICE NO.  -T#3	PRETEST  Y:
ORIFICE K FACTOR; -
                                                                     f 2 1n.H0
Orifice
manometer
reading
AH,
1n.H20
^o
Dry gas
meter
reading
vv
ft3
3bl3oo
&6.200
Temperatures
Ambient
°F
&q
^
Average
V
"F
*7
Dry gas meter
Inlet
VTif
°F
^3
^0
Outlet
w
°F
^
^^
Average
V
"F
n*
Duration
of
run
0
min.
^
Dry gas
meter
V ft3
/^/3^o
Vm
mstd'
ft3
&W
Vm
macf
ft3
/3.ZOO

Audit,
Y
OKI
Y
devia-
tion, %
-, 1 iL'2-
Audit
AHP,
in.H20
/•W

AH@ Devia-
tion, 1n.H20
o.o*.
 m
  std
            17.647(Vm)(Pbar. AH/13.6)

                                                  ft>
 m
   act
            1203( 0  )( K )(Pbar)
                (Ta + 460)
Audit
             m
             m
              std
                                                it Y
                                                     Audit Y
                                                                    x 100 - -
 Audit  AH@  =  (0.0317)(AH)(Pbar)(Tm + 460)
 Audit  Y  must  be  1n  the  range, pre-test Y ±0.05 Y.
 Audit  AH@ must be in  the  range pre-test AH@ ±0.15 inches
               Figure 3-10.  Mid-test audit report:   dry gas  meter
                      by critical orifice (Meter Box FB-1).
                                     3-15

-------
                       FIELD AUDIT  REPORT:  DRY GAS METER
                               BY CRITICAL ORIFICE
DATE:
                                         CLIENT:
                                                            - Q A vN
BAROMETRIC PRESSURE (Pbar):^7££ln.Hg  METER BOX NO.
ORIFICE NO. •   ?                        PRETEST
                                         AUDITOR*
ORIFICE K FACTOR:
"•- i. — ~ -f
Orifice
manometer
reading
AH,
1n.H20
J?,/O
Dry gas
meter
reading
vv
ft3
6 33.500
66?,6tfo
Temperatures
Ambient
Ta1/Taf
°F
3O
^O
Average
V
•F
T°
Dr
Inlet
T11/T1f
°F
tf/%6
Y
devia-
tion, %
-/.%
Audit
AH@,
in.H20
/•fz-
AH@ Devia-
tion, 1n.H20
O .0-2-
 m
  std
            17.647(Vm)(Pbar + AH/13.6)
 m
   act
            1203( 0 )( K )(Pbflr)
               (Ta * 460)
                         172
                                                  ft'
 Audit  Y
                                deviation
                                             Aud1t Y
              "std
                                                       S^^51 Y   x 100 - -/.
Audit  AH0 =  (0.0317)(AH)(Pbar)(Tm + 460)
 Audit Y must be  1n the  range, pre-test Y ±0.05 Y.
 Audit AHP must be in the  range pre-test AHP ±0.15 inches H,0.
                                                                            1n.H0
                Figure 3-11.  Mid-test audit report:  dry gas meter
                      hv rritical orifice (Meter Box FB-5).
                                     3-16

-------
                       FIELD AUDIT REPORT:   DRY  GAS METER
                               BY CRITICAL  ORIFICE
DATE:
                                         CLIENT:
BAROMETRIC PRESSURE (Pbar): P£~>Z-1n.Hg  METER BOX  NO.

ORIFICE NO. ^6                       PRETESTJfc-
                                                                          1n.H,0
ORIFICE K
Orifice
manometer
reading
AH,
1n.H20
P.*S
FACTOR: 5-P^X/O" * AUDITtfc^^W^f^^eW^-^

Dry gas
meter
reading
vv
ft3
1W.OOC
-tfO.tov
Temperatures
Ambient
Ta1/Taf
°F
6>3
^
Average
Ta-
"F
^
Dr
Inlet
W
°F
^^
•fc^
/ gas meter
Outlet
Toi/Tof»
°F
«^4
-S^r-
Average
V
•F
^.Z5
Duration
of
run
min.
/**
Dry gas
meter
V fts
/35oo
Vm
mstd«
ft3
/PdM.
y
act*
ft3
/P.?V9-

Audit.
Y
^.-w&
Y
devia-
tion, %
aVo
Audit
AHP,
1n.H20
/.-Z-2-

AH@ Devia-
tion, 1n.H20
o.of .
 m
  'std
            17.647(V  )(P.   + AH/13.6)
                    m
  "act
             1203( 0  )(  K  )(Pbar)
                (Ta  + 460)
                          1/2
 Audit  Y
             m
              'act
             m
Y deviation -
               std
                                                   Y
                                                                      10° --
 Audit  AHP  =  (°-0317)(AH)(pbar)(Tm  + 460)
 Audit  Y  must  be  1n  the  range,  pre-test  Y  ±0.05 Y.
 Audit  AHC must be in the range pre-test AHP  ±0.15  inches
              Figure 3-12.  Mid-test audit report:   dry gas  meter
                     by critical  orifice (Meter Box FB-8).
                                     3-17

-------
                       FIELD AUDIT  REPORT:  DRY GAS METER
                               BY CRITICAL ORIFICE
DATE:
                                         CLIENT:
BAROMETRIC PRESSURE (P.._):^5z.In.Hg  METER BOX NO.
ORIFICE
ORIFICE K FACTOR:
                /Z~
                      bar
                                         PRETEST

                                         AUDITOR:
in.H,0
Orifice
manometer
reading
AH.
1n.H20
tf°
Dry gas
meter
reading
Vvf.
ft3
4t&.*oo
^,£00
Temperatures
Ambient
Tai/Taf
°F
?o
TO
Average
Ta-
"F
yo
Dry gas meter
Inlet
°F
*??-
4t
Outlet
°F
?Z-
*?(
Average
Tm-
'F
f/-5
Duration
of
run
min.
^~
Dry gas
meter
V ft3
l?3a°
Xtd'
ft3
/3.o^?>
Xcf
ft3
//. vv^
Audit,
Y
O.fVf
Y
devia-
tion, %
-/.fo
Audit
AHP,
1n.H20
/?f
AH@ Devia-
tion, 1n.H20
0°1 .
 m
  'std
            17.647(Vm)(Pbar * AH/13.6)
                                        • p.053 ft3
 m
  act
            1203( 0 )( K )(Pbflr)
               (Ta + 460)
                         1/2
Audit Y
                              Y deviation
                                             Audit Y
              "std
Audit AH@ = (0.0317)(AH)(Pbar)(Tm + 460)
                                                                    x 100 - -
                                                        AH/13.6)
Audit Y must be 1n the range, pre-test Y ±0.05 Y.
Audit AHP must be in the range pre-test AHP ±0.15 inches H~0.
            Figure 3-13.   Mid-test audit report:   dry  gas meter
                   by critical  orifice (Meter Box  FB-10).
                                                                            1n.H0
                                    3-18

-------
                THERMOCOUPLE DIGITAL INDICATOR
                        AUDIT DATA  SHEET
Date
                      Indicator No.   PP-O

Test Point
No.
1
2
3
4
Millivolt
signal*




Equivalent
temperature,
•F*
3^
^00
-5VO
//f^
Digital Indicator
temperature reading,
•F
33
*2^
t52>?
J144
Difference,
X
*.*
*.*<
.1
0
Percent difference must be less than or equal  to 0.51.
Percent difference:
  (Equivalent  temperature °R- Digital Indicator temperature reading CR)(100£)
                             (Equivalent temperature °R)
Where °R • °F * 460°F
  These values are to be obtained from the calibration data sheet for the
  calibration device.
             Figure  3-14.  Mid-test thermocouple digital  indicator
                     audit data sheet (Indicator No. 220).
                                     3-19

-------
                THERMOCOUPLE DIGITAL INDICATOR

                        AUDIT DATA SHEET
Date
Indicator No.
/
                                                 Operator
Test Point
No.
1
2
3
4
Millivolt
signal*




Equivalent
temperature,
•F*
32-
poo
-5^0
//fV
Digital Indicator
temperature reading,
•F
P
-------
Audit Name:-z-





Deviation
/
2-
^
J>
o
2_





Max. Allowable
Deviation
5°F
5°F
2°F
7°F
See table
0.7%
0.5 grams
0.20 in. Hg
Reference temp. °F
Max. deviation °F
32-140
7
141-273
9
274-406
11
407-540
13
541-673
15
674-760
17
 * Correction factor:
     NWS value (in. Hg) - [Altitude (ft)/1000(ft/in.  Hg)] + 0.74 in.  Hg**
** 0.74 in. Hg is the nominal correction factor for the reference barometer
   against which the field barometer was calibrated.
If 1t is not feasible to perform the audit on any piece of equipment, record
"N/A" in the space provided for the data.

              Figure 3-16.  Mid-test onsite audit data sheet.
                                     3-21

-------
Audit Name:
                          ON-SITE AUDIT  DATA  SHEET
Date:
%c/

Equipment
^ / Meter box
T5>5 inlet thermo.
£o , Meter box
^fy outlet thermo.
^^ Impinger
436 thermometer
Stack
thermometer
or
Thermocouple
Orsat
analyzer
Trip
balance
Barometer
Reference
ASTM-3F at
ambient temp.
ASTM-3F at
ambient temp.
ASTM-3F at
ambient temp.
ASTM-3F at
ambient temp.
ASTM-3F at
stack temp.
% O^ 1n
ambient air
IOLM std.
weight
Corrected*
NWS value
Reference
Value
•57
1?
JJ


20.8%
fa
M
Value
Determined
li
S3
53
S-%
rr





Deviation
*~ 1
^>
o
1





Max. Allowable
Deviation
5°F
5°F
2°F
7°F
See table
0.7%
0.5 grams
0.20 in. Hg
Reference temp. °F
Max. deviation °F
32-140
7
141-273
9
274-406
11
407-540
13
541-673
15
674-760
17
 * Correction factor:
     NWS value (in. Hg) - [Altitude (ft)/1000(ft/in. Hg)] + 0.74 in.  Hg**
** 0.74 1n. Hg 1*the nominal correction factor for the reference barometer
   against which the field barometer was"calibrated.
If 1t is not feasible to perform the audit on any piece of equipment, record
"N/A" 1n the space provided for the data.

             Figure 3-17.   Mid-test  onsite  audit data sheet.
                                     3-22

-------
                      FIELD AUDIT REPORT:  DRY GAS METER
                              BY CRITICAL ORIFICE
                                        CLIENT:   U5£PA
DATE:       ^     	

BAROMETRIC PRESSURE (Pbar)= '¥Ll£ In.Hg  METER BOX NO.  F£ - I	
ORIFICE NO.    i3	^_  PRETEST Y: Q.^k^   AHP 1,72,  1n.HtO

ORIFICE K FACTOR:   6.377*1&"^      AUDITOR:
Orifice
manometer
reading
AH,
1n.H20
2.I5
Dry gas
meter
reading
YV
ft3
V9o,ozr
£83.330
Temperatures
Ambient
T.1/Taf
°f
78
-78
Average
V
°F
78
Dry gas meter
Inlet
W
°F
11
S^
Outlet
T0l/T0f
°F
11
13
Average
V
°F
%zr
Duration
of
run
0
min.
X5.O
Dry gas
meter
V ft3
m*
'3.J0T
mstd»
ft3
/if/0
Xcf
ft3
/2,m
Audit,
Y
tiro
Y
devia-
tion, X
-/.JT
Audit
AHP,
1n.H20
/,7T
AH(? Devia-
tion, 1n.H20
-i-o, 05-
 "std
                        460)
                                                 ft3
 m
  act
            1203(  0 )(  K )(Pbar)
               (T  * 460)
                         1/2
Audit Y =
             m
              'act
                              Y deviation
                                             Audit Y - Pre-test Y
                                                     Audit Y
x 100 *  -
             "std
Audit AH(? = (0.0317)(AH)(Pbar)(Tm + 460)
Audit Y must be 1n the range, pre-test Y ±0.05  Y.
Audit AHP must be In the range pre-test AHP ±0.15  Inches
            Figure 3-18.  Post-test audit report:   dry  gas meter by
                     critical  orifice (Meter Box  FB-1).
                                     3-23

-------
                       FIELD AUDIT REPORT:  DRY GAS HETER
                               BY CRITICAL ORIFICE
DATE:
                                        CLIENT:   (J5£Pf>
BAROMETRIC PRESSURE (P):  ffiA?1n.Hg  METER BOX NO.
ORIFICE NO.
ORIFICE K FACTOR:
PRETEST Y: p.983     AHP L

AUDITOR:
                                                                         1n.H20
                                                      no&
Orifice
manometer
reading
AH,
1n.H20
/3o
Dry gas
meter
reading
W
ft3
gg^.730
S
Average
V
"F
-78
Dry gas meter
Inlet
VTif
°F
&(*
&8
Outlet
T0l/T0f
°F
go
^2^
Average
Tm-
°F
H
Duration
of
run
0
min.
is-.o
Dry gas
meter
V ft'
II Mo
mstd»
ft3
n\n
Xcf
ff
IOXST
Audit,
Y
All
Y
devia-
tion, %
-ivf
Audit
AH0,
in.H20
/.
-------
                       FIELD AUDIT REPORT:  DRY GAS METER
                               BY CRITICAL ORIFICE
DATE:
                                      CLIENT:
BAROMETRIC PRESSURE (Pbar);^/O 1n.Hg  METER  BOX NO.  FB-ff	
ORIFICE NO.     1	               PRETEST  Y: Q.99O     AH& 1.91    1n.H20
ORIFICE K FACTOR:

Orifice
manometer
reading
AH,
1n.H20
yt\o
Dry gas
meter
reading
vv
ft3
oiinoo
£X/f.Zeo
Temperatures
Ambient
Ta1/Taf
°F
^
*3-2~
Average
Ta-
"F
PO
Dr
Inlet
°F
^^f
^^
/gas meter
Outlet
VTof
°F
^0
^3>
Average
°F
P5.r
Duration
of
run
0
min.
3f-J'
Dry gas
meter
V fts
M tioo
Vm
mstd'
ft3
Zf.ftK
Vm
"act'
ft3
Zy,2T7

Audit,
Y
,?rf
Y
devia-
tion, %
-,n
Audit
AHP,
in.H20
l.lf-

AH(? Devia-
tion, 1n.H20
-t-.o? .
            17.647(Vm)(P    .AH/13.6)
                    m
                        460)
'act
            1203( 0  )( K )(Pbflr)
                (Ta + 460)
                         1/2
                                                  ft3
 Audit Y •
             m
              'act
             m
                            Y deviation
               std
Audit Y - Pre-test Y
        Audit Y
x 100 = -
 Audit  AH?  =  (0.0317)(AH)(Pbar)(Tm + 460)
 Audit Y must be  1r,  the  range, pre-test Y ±0.05 Y.
 Audit AHP must be in  the  range pre-test AHP ±0.15 Inches H^O.


           Figure 3-20.   Post-test  audit report:  dry gas meter by
                        critical  orifice (Meter Box FB-8).
                                                                          1n.H20
                                    3-25

-------
                       FIELD AUDIT REPORT:  DRY GAS METER
                               BY  CRITICAL ORIFICE
DATE:
                                       CLIENT:
BAROMETRIC PRESSURE (Pbar):<2§/£ In.Hg  METER BOX NO.  £B~  (O	
ORIFICE NO. 	to        	 PRETEST YQ.96T7      AHP \.l\   1n.H,0
ORIFICE K FACTOR: 5.£*M * IcT^          AUDITOR:
1
Orifice
manometer
reading
AH,
1n.H20
2,3-0
Dry gas
meter
reading
W
?Z2 .Too
W7*°
Temperatures
Ambient
Tai/Taf
°F
"W
??
Average
Of
Dr
Inlet
T11/T1f
°F
*?*f
^
y gas meter
Outlet
°F
^0
^/
Average
Tm-
"F
ttr
Duration
of
run
0
min.
/f^
Dry gas
meter
V ft3
a/^
Vm
mstd*
ft3
Iti*?
Vm
macf
ft3
/^JT7

Audit,
Y
,1U
Y
devia-
tion, %
-,lo
Audit
AHP,
1n.H20
/,fr

AH(? Devia-
tion, 1n.H20
^,^/ .
"std
            17.647(Vm)(Pbar + AH/13.6)
                        460)
                                                  ft3
 m
   act
            1203( 0  )( K )(Pbflr)
                (T  * 460)
                        172
                                                 t3
 Audit  Y =
             m
              'act
             m
                            Y deviation
Audit Y - Pre-test Y
        Audit Y
x 100 «  ^
              'std
 Audit  AHC<  =  (0.0317)(AH)(Pbar)(Tm + 460)
 Audit Y must  be  1n  the  range, pre-test Y ±0.05 Y.
 Audit AH3 must be 1n  the  range pre-test AHP ±0.15 Inches H20.


            Figure 3-21.   Post-test audit report:   dry gas meter by
                        critical orifice (Meter Box FB-10).
                                                                            i.HoO
                                    3-26

-------
                THERMOCOUPLE DIGITAL INDICATOR
                        AUDIT DATA SHEET
Date
Indicator No. IIP
                                                 Operator
Test Point
No.
1
2
3
4
Millivolt
signal*




Equivalent
temperature,
•F*
•*>?•
^oo
-s*t°
//
2°?^
-5*5 f
//W
Difference,
1
,w
.v>
.or
.00
Percent difference must be less than or equal  to  0.5%.

Percent difference:

  (Equivalent temperature °R- Digital Indicator  temperature reading eR)(100t)
                             (Equivalent temperature  °R)
Where °R « °F * 460°F
  These values  are to be obtained from the calibration data sheet for the
  calibration device.
           Figure 3-22.   Post-test thermocouple digital indicator
                    audit data sheet (Indicator No.  220).
                                      3-27

-------
                THERMOCOUPLE DIGITAL INDICATOR
                        AUDIT DATA SHEET
Date
- /f "
Indicator No.   2,21
                                                 Operator
Test Point
No.
1
2
3
4
Millivolt
signal*




Equivalent
temperature.
•F*
32.
0
-SfO
/PW
Digital Indicator
temperature reading,
•F
*
-------
                     FIELD AUDIT REPORT:  DRY GAS METER
                            BY CRITICAL  ORIFICE
DATE: ---
                       CLIENT:
BAROMETRIC PRESSURE (P):P#£? 1n.Hg  METER BOX NO. f&
ORIFICE NO.
ORIFICE K FACTOR ;
                    bar
                       PRETEST
                       AUDITOR
                                                            /,?31n.H.O
                                                 |"V^57^SE*X*^^^
Orifice
manometer
reading
AH,
1n.H20
/.io
Dry gas
meter
reading
w
ft3
(0?4.*>*. 300
Temperatures
Ambient
Ta1/Taf
°F
£,*
6K
Average
°F
^
Dry gas meter
Inlet
°F
To
?-o
Outlet
VTof
°F
^>?
(o^
Average
°F
&.*
Duration
'Of
irun
- 0
min.
IU> 6O
Dry gas
 meter
V fts
"'std'
ft3
         m
                        'act'
                       ft3
                             offz
 AH@ Devia-
tion, 1n.H20
 m
  'std
  'act
Audit Y
           17.647(Vm)(Pbar * AH/13.6)
                        **"
           1203(  0  )( K )(Pbflr)
              (Tfl  + 460)
         1/2
            m
             'act
              Y deviation
            m
             'std
                                         Audit Y - Pre-test Y
                                                Audit Y
  100
Audit AH(3 =  (0.0317)(AH)(Pbar)(Tm + 460)
 Audit Y must be 1n the range, pre-test Y  ±0.05 Y.
 Audit AHP must be in the range pre-test AH@ ±0.15  inches H^O.


         Figure 3-24.  Example of unacceptable dry gas meter audit.


                                 3-29

-------
     PEI personnel calculated the sampling rates on site.  The
data were rechecked and validated at the end of the test program
by computer programming.  Some minor discrepancies between the
hand calculations and computer printouts resulted primarily
because of round-off error.  Overall, the data compared favor-
ably.  Figure 3-25 presents an example calculation form PEI using
during this test program.  Computerized example calculations are
presented in Appendix A.
     As an additional check of the reliability of the method used
to analyze the samples/ two blank trains were assembled in the
recovery area, capped off, and set aside for about 2 hours.  The
first blank train was assembled at the beginning of the test
series using clean glassware.  On the same day as Quad Runs 1, 8,
and 9  (EMSL work), the second blank train was assembled with
glassware used during previous sampling runs.  The blank trains
were recovered in the same manner as the test samples.  These
samples were shipped to the laboratory and analyzed by the same
procedures as those used for the actual emission samples.  In
addition to the blank sampling train, aliquots of the field
reagents used in the collection and recovery of the samples were
obtained daily and analyzed by the same procedures as those used
for the actual samples.  Table 3-2 presents the results of the
blank sample trains and field blank analyses.  The results are
considered reasonable and indicate that background arsenic
contamination was not a problem in the sample recovery area.  The
results of the blanks are relatively small.
                               3-30

-------
                              ISOKINETIC CALCULATION
                                                41.
SITE
. TEST NO.

1. »ol*» Of dry fit M^lod COrrvcUd to
tUnd«r« condition!. toU: V Milt ta
corrtcttd for iMkigt 1f in/ llikigt
nUt tacttd L ).
r, „, i
t_ • 11. «s » v • T tar • TTT
•ltd • I !• J
1$^ ^
V.<«
2. »elu«» of Mtf Mpor it tUnferd COn-
•itlont, ft'.
V • 0.04707»j •
"ltd IC
3. Neiltvrt content 1n tuck 911.
1 • "ltd •
•ltd "ltd
4. Or; acltculir «»1JM of lUCk fit.
Mfl • 0.440 (t COj) • O.UO (t Oj)
• 0.7(0 (» «j • » CO) •
I. ltol»cul«r Might of tuck tu-
ft. Stick vtlocU/ It tUCt condition*.
*pt


T. iMttMtlC Mrt«t««n
». T.
S 1 • "tU 1*1 1T.V
t. * B. • • I ». • (!••_.)
• • 9 VI

V ft*
t
%.r. 1".Hg
AH. 1n.H20
T«-'R
V . dscf
"std
V1c. 9
vw .ft3
*$td
Bws
J-Bws
* C02
%JXf^).
1 N2 * X CO
Md, Ib/lb-nole
H$. Ib/lbHBlt
P 1fl t4
r»tit1c* in'H2
P,. 1n.H9
T^. "R
ysr
Cp
V,. fps
On. 1n.
9, «1n.
S I
RUN 1
ZJ.177
rftf
W/f
• 9c3
i
3*,o?1
-73.Y
3 Ml
,cd
.W
y.!>
<^n
1$.
£-?,?-
&#A
-.£$,
&,<*<{
to&
*•* */ 0
6y
f/3 "^_
,2*7

fOt.T>
RUN 2
v/.r^3
.983-
Wi
/
T^>/
y..7
,^V
AOL





















RUN 3
Vfd-5
,^7
W.tf
l.'f
**1
«,. ,'73
r/y
3-*3
.CSV
.?/6
--™

«-=



«BK^
^^c>;
.ffo
^?yr
7-.ll
Si\
f*.7V
43.4
y.vo
.^77
,fc3
	 ^
^^
^>

"^
4p-
k
I ^/7 T./












*•* • ' I
^^^^^




,^f

•/f,^
.ifi

f^.O.
         Figure 3-25.   Example of onsite  calibration data  sheet.
                                      3-31

-------
                         TABLE  3-2.  ARSENIC BLANK DATA
                      Blank  sample  train arsenic values'
Train No.
1
2
Filter, yg
32.8
30.8
NaOH probe
rinse, yg
8.0
7.8
Impinger
section, yg
22.2
11.6
Total train
blank, yg
63.0
50.2
                          Field  blank  arsenic values
Date
samples
taken
5/17/84
5/18/84
5/19/84
Corresponding
Run No.
10 + 11
12 + 13
CD1 - CDS
Average blank values
Filter
total , yg
26.7
29.3
28.8
28.3
NaOHb ,
mg/liter
0.0168
0.0137
0.0153
0.0153
H20C,
mg/liter
0.0085
0.0101
0.0111
0.0099
 Sample  train  was  fully  assembled  in  recovery  area and  then  recovered and
 analyzed  as a sample.

5Between 235 and 238  ml 'of  NaOH was used  to  rinse the probe.  Between 179
 and 456 ml of the NaOH  was used to rinse Impingers  1 and  2.  Between 128
 and 184 ml of the NaOH  was used to rinse Impingers  3 and  4.  Between 119
 and 180 ml of the NaOH  was used to rinse the  connector.   The maximum blank
 for the NaOH  corresponds to 6 yg  for the probe  rinse,  8 yg  for  the  impinger
 samples,  and  3 yg for the  connector  samples.

"On  both days, 300 ml of water was added  to  arsenic  Impingers 1  and  2 and
 150 ml  to Impingers  3 and  4.  The maximum blank for the water corresponds
 to  3 yg for Impingers 1 and 2 and 2  vg for  Impingers 3 and  4.
                                    3-32

-------
     Laboratory reagent blank analyses were performed during the



analysis of the field samples.  The results of these analyses are



presented in Table 3-3.  The average value for five filter blanks



was 29.7 yg out of a range of 26.7 to 32.8; a blank correction of



30 yg was used to correct all the reported data.  All of the



blank values for the rinse and impinger samples were near the



analytical detection limit of 2 to 8 yg.  Because of the vari-



ability and relatively small value of the blank, no average value



was determined and no blank corrections applied.



     Values below 50 yg were considered insignificant and not



reported because 8360 yg  (7570 on the filter) was the minimum



amount of arsenic determined in any train and the blanks for the



liquid sample fractions varied considerably.



     Each sample was first analyzed by the flame technique.



Samples whose concentrations were below 30 mg/liter were also



analyzed using the graphite furnace.  Actual sample concentra-



tions were either gre.ater than 100 or less than 10 ppm.  The



30-mg/liter limit was based on previous experience with Method



108, which indicated good agreement above this level.  As the



analyses were completed and the data were reduced by the labora-



tory, the results were reviewed by the Quality Assurance Officer



(QAO).  The QAO reviewed instrument calibration, the analysis of



the standard reference solution  (SRS), agreement between flame



and furnace results, and general consistency of the data.  He



then prepared a list of samples for reanalysis.
                               3-33

-------
           TABLE 3-3.   ARSENIC LABORATORY  REAGENT  BLANK  DATA
Date
(1984)
6/8
Filter
total , yg
2.4
Rinse,3
mg/ liter
0.0079
Impingers,
mg/liter
0.0079
Connector,0
mg/ liter
0.0079
 Between 235 and 328 ml  of sample were received  as  the  rinse fraction.
 The maximum laboratory  reagent blank corresponds to  3  yg  for  this
 fraction.

 Between 532 and 817 ml  of sample were received  as  the  Impingers  1  and
 2 fractions and between 280 and 340 ml  as the  Impingers 3 and 4  frac-
 tion.  These correspond to maximum laboratory  reagent  blanks  of  6  yg
 and 3 ug,  respectively.

cBetween 119 and 180 ml  of sample were received  as  the  connector
 fraction.   The maximum  laboratory reagent blank corresponds to 1 yg for
 this fraction.
                                 3-34

-------
     The analysis was performed in five batches by flame atomic



absorption.  Eighteen sets of standards (0, 10, 30, 50, 80, 100



ppm) were analyzed with the samples.  The linear regression data



for all the standards analyzed with a given batch of samples are



presented in Table 3-4.  The average correlation coefficient is



0.9989, out of a range of 0.9994 to 0.9985.  The average detec-



tion limit is 2.3 ppm.  A value of twice the range of the 0-ppm



standard above the Y-intercept was used to calculate the detec-



tion limit.  A standard reference solution independently prepared



from AS-O., with a nominal value of 150 ppm was analyzed  (1-2



dilution) with each set of standards.   (Standards were prepared



from a commercially available 1000-ppm standard solution.)  The



average value obtained in the 18 analyses of this standard



reference solution (SRS) was 157.4 ppm, with a standard deviation



(SD) of 3.81 ppm  [2.4 percent relative standard deviation  (RSD)].



Only 1 of the 18 determinations made fell outside the range of



the mean ±2 SD  (one was 166 ppm).



     These data indicate that the precision and accuracy of the



flame atomic absorption analyses are well within acceptable



limits.  The percent difference of the average measured value of



the SRS and its predicted value is 4.9 percent; the RSD of the



measured value is 2.4 percent.



     The results of the audit samples supplied by EPA and deter-



mined by flame atomic absorption  (listed in Table 3-5) are



consistent with the data just presented.  The relatively large



difference at 10 ppm is predictable in that it is only 5 times



the average detection limit.




                                3-35

-------
TABLE 3-4.   LINEAR  REGRESSION DATA (FLAME)

Date
(1984)
6/4
6/5
6/5
6/11
6/13
No. of
standard
curves
4
2
2
5
5


Y-intercept
0.0031
0.0048
0.0056
-0.0011
0.0041


Slope
0.00489
0.00500
0.00496
0.00490
0.00450

Correlation
coefficient
0.9990
0.9994
0.9985
0.9990
0.9987

Detection
limit, ppm
2.1
1.2
1.6
1.6
4.9
                    3-36

-------
                       TABLE  3-5.  ARSENIC AUDIT RESULTS
EPA No.
B-3-I
B-4-I
G-l-I
G-3-I
H-l-I
H-2-I
B-3-I*
B-4-I3
WP-4753
Cone 6
Lab No.
DC329
DC330
DC331
DC332
DC333
DC334
DC329
DC330
DM562
Volume, ml
500
500
500
500
500
500
500
500
1000
Arsenic
concentration
EPA values,
mg As/liter
10
10
100
100
40
40
10
10
0.207-0.393
Measured,
mg/ liter
10.2
11.7
111
107
43.9
43.9
11.1
10.7
0.356
Total
As, mg
5.12
5.83
55.6
53.3
22.0
22.0
5.54
5.34
0.356
Graphite furnace analysis.
                                   3-37

-------
     Table 3-6 presents the results of 10 samples checked by the

method of standard addition.  The slopes of all the standard

addition analyses are between 0.9 and 1.1 except for those of

DM697, DM688, and DM660, which is probably due to an error in the

spiking or the fact that no given point was in the regression

analysis because a less-than value is unusable.  An analysis of

the results of the unspiked samples and the X-intercepts  (stan-

dard addition values) revealed that only Sample DM697 showed a

significant difference.  The results for DM697 were expected,

based on the slope; the results of standard addition show no
                                                                \
consistent bias attributable to the sample matrices.

     The samples were analyzed by atomic absorption in which

graphite furnace techniques were used.  All samples below 30 ppm

were analyzed by furnace techniques.  Sample concentrations were

either greater than 100 ppm or less than 10 ppm.  Values obtained

from flame and furnace techniques cannot be accurately compared

below 10 ppm because this value is too close to the flame detec-

tion limit.  Twelve sets of standards (0, 0.01, 0.05, 0.10, and

0.15 mg/liter) were analyzed with the samples.  All the data sets

were reduced by linear regression analysis  (see Table 3-7).  The

average correlation coefficient for the linear regression anal-

ysis was 0.9970, out of a range of 0.9980 to 0.9954.  The average

detection limit for the graphite furnace was 0.0033 ppm.  A value

of twice the range of the 0-ppm standard above the Y-intercept

was used to calculate the detection limit.
                               3-38

-------
                 TABLE 3-6.  ARSENIC STANDARD ADDITION  RESULTS
Lab Number
DM643 Filter



DM670 Filter



DM697 Filter



DM736 Filter



DM637 Bomb



DM688 Bomb



DM650 Rinse



Spike,
ppm
0
20
30
40
0
20
30
40
0
20
30
40
0
20
30
40
0
20
30
40
0
20
30
40
0
20
30
40
Previously
determined
flame, ppm

26.16



30.07



31.45



28.24



<2.6



<3.5



<2.9


Measured,
ppm
22.45
45.58
52.03
62.70
29.12
47.80
56.92
69.60
35.35
48.47
57.37
66.26
26.90
47.36
60.48
66.93
<4.9
10.88
18.67
29.79
<4.9
11.55
20.22
28.45
<4.9
8.44
18.89
27.79
Linear
regression analysis
Slope = 0.992
Y intercept = 23.36
Corr. = 0.9955
X intercept = -23.54
Slope = 0.995
Y intercept = 28.48
Corr. = 0.9974
X intercept = -28.64
Slope = 0.769
Y intercept = 34.55
Corr. = 0.9964
X intercept = -44.90
Slope = 1.030
Y intercept = 27.24
Corr. = 0.9957
X intercept = -26.45
Slope = 0.946
Y intercept = 0.87
Corr. = 0.9949
X intercept = -0.92
Slope = 0.845
Y intercept = 3.2
Corr. = 0.9999
X intercept = -3.76
Slope = 0.968
Y intercept =0.98
Corr. = 0.9989
X intercept = -1.01
(continued)
                                    3-39

-------
TABLE 3-6  (continued)
Lab Number
DM740 Rinse



DM660 Impinger



DM717 Impinger



Spike,
ppm
0
20
30
40
0
20
30
40
0
20
30
40
Previously
determined
flame, ppm

5.5



<2.1



<1.3


Measured,
ppm
<4.9
10.66
19.11
30.45
<4.9
11.55
17.11
26.45
<4.90
7.55
16.00
25.56
Linear
regression analysis
Slope = 0.990
Y intercept =0.28
Corr. = 0.9965
X intercept = -0.28
Slope = 0.745
Y intercept = 3.47
Corr. = 0.9894
X intercept = -4.66
Slope = 0.900
Y intercept =1.64
Corr. = 0.9994
X intercept = -1.82
                                   3-40

-------
TABLE 3-7.   LINEAR  REGRESSION DATA (FURNACE)
Date
(1984)
6/8
6/11
6/15
No. of
standard
curves
2
4
4
Y-intercept
-0.0011
0.0019
-0.0014
Slope
4.081
4.316
3.853
Correlation
coefficient
0.9980
0.9975
0.9954
Detection
limit, ppm
0.0039
0.0028
0.0031
                  3-41

-------
     A standard reference solution independently prepared from

As203 with a nominal value of 0.0750 ppm was analyzed with each

set of standards.   (Standards were prepared from a commercially

available 1000-ppm standard solution.)  The average value ob-

tained for the 21 analyses of this SRS was 0.0751 ppm with a

standard deviation of 0.0027  (3.6 percent relative standard

deviation).  Historically, the mean value for this SRS is 0.0762,

with a standard deviation of 0.0027.  The values obtained for the

SRS solution during this project are in good agreement with our

historical data.  These data indicate that the precision and

accuracy of the furnace atomic absorption analyses are well

within acceptable limits.  The difference in the average measured

value of the SRS and its predicted value is 0.2 percent; the RSD

of the measured value is 3.6 percent.

     The results of audit samples analyzed using the graphite

furnace were listed in Table 3-5.  These values are consistent

with the previous value and the accepted values.  The results of

duplicate analysis are presented in Table 3-8.  The absolute

value of the percent difference was calculated according to the

following equation.


                          X, _ X2
          % Difference = 	
          where X, and x~ are the individual values

                X is the average of X., and X-
                               3-42

-------
                 TABLE 3-8.  DUPLICATE ANALYSIS DATA
Sample fraction
Filter







Bomb
Rinse




Impinger

Filter3
Bomb3



Rinse6

Impinger







Arsenic, g
8,310, 8,470
8,770, 9,130
9,690, 9,590
10,900, 11,100
10,900, 11,300
10,400, 10,300
10,800, 10,700
10,900, 10,400
99, 93
351, 294
498, 467
290, 269
610, 648
768, 657
50, 48
4, 5
36, 35
39, 32
61, 56
40, 28
66, 51
678, 693
623, 629
35, 30
32, 29
4, 49
227, 191
33, 30
5, 45
43, 39
153, 55
% Difference
1.4
3.0
1.0
1.8
3.6
1.0
0.9
4.7
6.2
17.7
6.4
7.5
6.0
15.6
4.1
22.2
2.8
19.7
8.5
35.3
25.6
2.2
1.0
15.4
9.8
170
17.2
9.5
160
9.8
94
 Same  aliquot  analyzed on different days.
3Sample  aliquot  prepared and analyzed on different days.
                               3-43

-------
     Filter samples were analyzed using flame atomic absorption,
and all other sample fractions were analyzed using the graphite
furnace technique.
     The first 16 values reported are based on duplicate analysis
of the same sample aliquot on the same day using the same cali-
bration curves.  The agreement on the front filters is very good.
These filters contained better than 90 percent of the total
arsenic collected.  The average percent difference for the
primary filter is 2.4 percent.  The agreement for the other
sample fractions is acceptable and will not have a significant
influence on the overall method precision because they represent
less than 10 percent of the total arsenic collected.
     The next five values reported are based on repeat analysis
of the same sample aliquot on different days using different
calibration curves.  The agreement is good considering the
relatively small amount of arsenic contained in these fractions.
     The last 10 values reported are based on repeat analysis of
different sample aliquots prepared and analyzed on different days
using different calibration curves.  The agreement for the two
rinse samples  (the only two containing a significant amount of
arsenic) is very good.
                              3-44

-------
                            SECTION 4


               SAMPLING LOCATION AND TEST METHODS



     This section describes the sampling sites and test methods


used to characterize arsenic emissions from each source evalu-


ated.


     A four-train (quad) sampling system was used to collect


samples at the exit stack of the glass melting furnace.  This


system allows four trains to sample simultaneously at essentially


a single point in the stack (see Figures 4-1 and 4-2).


     Because this sampling approach allows four trains to sample


simultaneously at essentially a single point, it reduces the


effect of variations in the velocity and particulate profiles on


the sampling results.  It also permits a statistically signifi-


cant number of samples to be taken in a short amount of time.


Further, since all four trains are identical for every run, the


within-train precision can be determined at the same time as the


relationship of the different trains is being compared.  This


methodology for determining method precision was developed and


validated in a previous EPA study.*  A total of four quad-train


runs representing 16 individual samples were collected.  During
*
 Mitchell, W. J., and M. R. Midgett.  A Means to Evaluate the
 Performance of Stationary Source Test Methods.  ES and T,
 10:85-88, 1976.
                               4-1

-------
BACKUP
METHOD 5
FILTER
                                   IMPINGER
                                    TRAINS
                       (204°C)                   (288°C)
                                   HEAT BOXES
              BACKUP
             •METHOD  5
              FILTER
              (izrc)
                                     B A D C
                                     CO CD CD CO
                                     oo oo
                                     oc cc oc o;
                                     0.0. 0. O-
oD  oA

OC  oB
                                                        BACK VIEW

      Figure 4-1.   Quad  train  system  for elevated temperature tests

                                   4-2

-------
  3 1n.
                                           THERMOCOUPLE
    2 1n.

1 in/

1 1n.


„,*] '
1n. H
3/4 1n
S" TYPE PITOT TUBEft
         2:	
          [*     2 In. -—+\
Figure 4-2.   Four-train sampling system showing  nozzle, pitot

            tube, and thermocouple position.
                            4-3

-------
these runs, a single Method 108 train was run with the nozzle


positioned as close to the quad nozzle arrangement as possible


without causing interference.


     All samples were collected at the furnace exit stack as


depicted in Figures 4-3 and 4-4.  Ambient ejector air is con-


trolled automatically to maintain furnace pressure and is intro-


duced angularly as depicted in Figure 4-3 at a volume ratio of


about 1:1 to the furnace gases.  According to plant personnel,


the furnace gas temperature is about 760°C (1400°F),  and exit gas


temperatures ranged between 260° and 316°C (500° and 600°F),


which indicated that the gases were relatively well mixed at the


sample cross section.  Single-point, isokinetic sampling tech-


niques were used in each quad run and reference train tests.


Prior to sampling, a complete velocity and temperature profile


was established using procedures described in EPA Methods 1 and


2.*  The velocity and temperature data were used to select sample


nozzle sizes so as to maintain isokinetic sample rates and ensure


adequate sample volume [0.85 dscm (30 dscf)] in each train.  The


quad nozzle assembly was positioned approximately  52 cm  (20.5


in.) from the inside wall of the stack in each run.  Sampling


rates generally ranged between 0.014 dsm3/m  (0.50 dscf/m) and


0.017 dsm3  (0.60 dscf/m), and sampling times were typically 60


and 70 minutes.


     In the Method 108 traverse tests, 24 sampling points were


used to traverse the cross-sectional area of the stack.  Each
*
 40 CFR 60, Appendix A, Reference Methods 1 and 2, July 1983,


                               4-4

-------
                     142 on
       ,    .
       (3 ft 0 In
                                     BUTTERFLY
                                       VALVE
                                          CHAIN
                                        ADJUSTMENT
      7.6 m
       .       .
   (25 ft 0 1n.)
          1.5 in
      (5 ft 0 1n.)
                      (36 1n.)
                      V i.d. '
VENTURI
MIXING
   7.9 m
(26 ft 0 1n.)
1 '
0.76 m
(2 ft 6 1n
1.4 m
(4 ft 7 1n.]
...I I
._,
.) / EJECTOR \
l'"°\
CHAMBER
TRANSITION
f
EJECTOR
(AMBIENT
AIR)
r r
i \
SASES \
1 | FURNACE |
( *1 ft 6 1nJ
1 1

1

GROU
Figure 4-3.   Furnace exit  stack  elevation (no  scale)
                           4-5

-------
       3-1/2 1n. 1.d.
       SAMPLE PORT
                                              BUTTERFLY
                                                VALVE
                                                AXIS
     6 in.  i.d.
     SAMPLE PORT
     (QUAD  TRAIN)
STACK I.D. (3 SAMPLE
PORT: 104 cm (41 in.)
              CROSS SECTION
Figure 4-4.   Furnace exit stack sampling  port  location
                  (no scale).
                       4-6

-------
point was sampled for 2.5 minutes, yielding a total test time of


60 minutes.



4.1  SAMPLING AND ANALYTICAL PROCEDURES


     The sampling and analytical procedures used in this test


program followed those described in EPA Reference Methods 1


through 4* and proposed Method 108 as detailed in the site test


plan prepared by PEI and reviewed by EMB.  The procedures, which


are described briefly here, are detailed in Appendices D and F.


4.1.1  Velocity and Gas Temperature


     A Type-S pitot tube and an inclined draft gauge manometer


were used to measure gas velocity pressures at the test sites.


Temperature was measured with a thermocouple and digital readout.


During each sample run, velocity and temperature measurements


were taken at a single sampling point in the duct.  Prior to each


test series, separate velocity measurements were taken by tra-


versing the entire sample cross-sectional area to determine an


average value.  Measurements were taken in accordance with proce-


dures outlined in Reference Method 2 of the Federal Register.*


4.1.2  Molecular Weight


     Flue gas composition was determined in accordance with the


basic procedures described in Reference Method 3.*  Grab samples


were collected before any sampling began to establish baseline


contents of oxygen, carbon dioxide, and carbon monoxide.  Bag
*
 40 CFR 60, Appendix A, Reference Methods 1 through 4, July 1983.
                                4-7

-------
samples were collected at least twice daily during sampling at

each source and analyzed with an Orsat gas analyzer.  An inte-

grated bag sample was also collected during each Method 108

traverse test and analyzed using an Orsat gas analyzer.

     Method 108* was used to measure arsenic concentration except

that the impingers containing hydrogen peroxide  (H-O.) for SO-

determination were eliminated due to low  (less than 30 ppm)

concentrations of S0_.  All tests were conducted isokinetically

by regulating the sample flow rate relative to the gas velocity

in the stack as measured by the pitot tube and thermocouple

attached to the quad probe arrangement (see Figure 4-2).  Each

individual sampling train consisted of a heated glass-lined

probe, a heated 7.6-cm  (3-in.) diameter glass fiber filter  (What-

man Reeve Angel 934AH), and a series of four Greenburg-Smith

impingers followed by a vacuum line, vacuum gauge, leak-free

vacuum pump, dry gas meter, thermometers, and a calibrated ori-

fice.  In each train, probe and filter temperatures were moni-

tored using digital indicators and thermocouple leads  located in

each probe and immediately behind the Method 108 filter frit.  In

Quad Runs 10, 12, and 13, a 53-cm  (21-in.) glass connector was

used to attach the front filter to a backup filter maintained at

approximately 121°C.  The impingers followed the backup filter in

these runs.

     The amount of water collected in the impinger section of the

sampling train was measured gravimetrically at the end of each
 Method 108 is proposed.  40 CFR 61, Appendix B, Method 108,
 July 1983.


                               4-8

-------
sample run to determine the moisture content of the flue gas.



The contents of the first two impingers, each of which had been



charged initially with 150 ml of distilled water, were trans-



ferred to a polyethylene container.  These impingers and all



connecting glassware  (including the back half of the filter



holder) as well as a third (empty) impinger were rinsed with 0.1



N NaOH; the rinse was then added to the container.  The contents



of the first two impingers and 0.1 N NaOH rinse were analyzed for



arsenic by atomic absorption.  In the elevated temperature runs,



the third and fourth impingers were recovered and analyzed simi-



lar to Impingers 1 and 2.
                               4-9

-------
                            SECTION 5



                        PROCESS OPERATION






     Tests were performed on the uncontrolled emissions from a



regenerative natural-gas-fired glass melting furnace.  The fur-



nace evaluated has a pull-rate capacity of 90 to 100 tons/day and



produces primarily crystal glass utilizing arsenic as a condition-



ing and refining agent.  Furnace pressure is maintained by use of



an induced-draft ejector system as described in Section 4.  Data



collected during this study indicate that the furnace gases and



the ejector air were adequately mixed at the sampling location.



     Personnel from Radian Corporation  (an EPA contractor) moni-



tored the furnace operation during each test.  Appendix F of this



report contains a detailed process description and a summary of



furnace operating data.
                               5-1

-------