United States
Environmental Protection
Agency
Office of Air Quality
Planning and Standards
Research Triangle Park NC 27711
EMB Report 84-IBR-24
August 1984
Air
Industrial Boilers
Emission Test
Report

General  Electric Noryl
Products Division
Selkirk, New York

-------
             EMISSION TEST REPORT

      METHOD DEVELOPMENT AND TESTING FOR
        INDUSTRIAL BOILERS, PM AND NO
           General Electric Company
               Selkirk,  New York

                ESED NO. 76/13
               EMB NO. 84-IBR-24
                     by

             PEI Associates, Inc.
             11499 Chester Road
               P.O. Box 46100
        Cincinnati, Ohio  45246-0100
            Contract No.  68-02-3849
         Work Assignment Nos.  7 and 8
             PN 3615-7 and 3615-8
                 Prepared for

             Mr. Dennis Holzschuh
                 Task Manager
     U.S. ENVIRONMENTAL PROTECTION AGENCY
  EMISSION STANDARDS AND ENGINEERING DIVISION
          EMISSION MEASUREMENT BRANCH
RESEARCH TRIANGLE PARK, NORTH CAROLINA  27711
                 January 1985

-------
                                  CONTENTS
Figures                                                               iii
Tables                                                                 iv
Acknowledgment                                                         vi
Quality Assurance Element Finder                                      vii

1.   Introduction                                                       1

2.   Summary of Test Results                                            3

     2.1  Test protocol                                                 3
     2.2  Continuous emission monitor data                              6
     2.3  Particulate and fuel analysis test results                   14

3.   Quality Assurance                                                 30

     3.1  Continuous emission monitors                                 31
     3.2  Manual tests—particulate and NO                             51
                                          /\

4.   Sampling Locations and Test Methods                               64

     4.1  Sampling locations                                           64
     4.2  Continuous emission monitors—sample extraction,
           analysis, and data reduction                                64
     4.3  Particulate test methods and analytical procedures           69
     4.4  Manual test methods for NO                                   69
                                    7\

5.   Process Description and Operation                                 71

     5.1  Boiler description                                           71
     5.2  Burner description                                           73
     5.3  Operating history                                            74
     5.4  Control procedures                                           76

References                                                             78

Appendices

 A   Computer Printouts and Example Calculations                      A-l
 B   Field Data Sheets                                                B-l
 C   Laboratory Data Sheets                                           C-l
 D   Sampling and Analytical Procedures                               D-l
 E   Equipment Calibration Procedures and Results                     E-l
 F   Quality Assurance Summary                                        F-l
 G   Project Participants and Field Log                               G-l

                                      ii

-------
                                   FIGURES

Number                                                                Page
  1       Example NO  Calibration Curve                                40
                    A

  2       Example Op Calibration Curve                                 41

  3       Example CO Calibration Curve                                 42

  4       Example C02 Calibration Curve                                43

  5       On-Site Audit Data Sheet                                     53

  6       Field Audit of Dry Gas Meter (Meter Box FB-4)                 54

  7       Field Audit of Dry Gas Meter (Meter Box FB-8)                 55

  8       Example of an Unacceptable Meter Box Audit                   56

  9       Example Calculation Form Used by PEI During Test Series      57

 10       Audit Report NO  Analysis Results on Samples Received
           8/15/84       x                                             60

 11       Audit Report NO  Analysis Results on Samples Received
           8/16/84       x                                             61

 12       Audit Report NO  Analysis Results on Samples Received
           8/17/84 and 8/18/84                                         62

 13       Audit Report NO  Analysis Results on Samples Received
           8/22/84       x                                             63

 14       Boiler 4 Exit Stack (No Scale)                               65

 15       Boiler 5 Exit Stack (No Scale)                               66

 16       CEM System Layout                                            67

 17       Unit 5 Layout                                                72

 18       Coen Parallel Flow Type LEA Burner                           75

-------
                                   TABLES

Number                                                                Page

  1       Summary of Boiler Operating Parameters  and  Emissions  Tests   4

  2       Summary of Continuous Emission Monitoring Data               7

  3       Summary of Maximum and Minimum CEM Data by  Test  Block       10

  4       Relationship Between NO  Concentration  and  Excess  Air
           Levels—Residual Oil  x                                    11

  5       Relationship Between NO  Concentration  and  Excess  Air
           Levels—Natural  Gas                                        12

  6       Summary of NO  Emission Rates                               15
                       X
  7       Summary of Particulate Sampling Conditions, Boiler No.  5
           Stack                                                      17

  8       Summary of Particulate Emission Results                     18

  9       Between-Run Statistical Data for Similar Sample  Types
           and Boiler Loads                                           21

 10       Summary of Fuel  Analytical  Results                          22

 11       Method 5B Relative Percent  Weight Loss  at  160°C              26

 12       Residual Sulfate (S0.=) Analysis of Methods 5 and  5B
           Sample Fractions                                           27

 13       Analytical Results From Ether/Chloroform Extraction of
           Back-Half Solutions                                        29

 14       Monitor Stratification Test—Boiler 4 (8/14/84)              32

 15       Monitor Stratification Test—Boiler 5 (8/16/84)              32

 16       Test Results for NO  Monitor 24-Hour Zero  and Calibration
           Drift             x                                        33

 17       Test Results for CO Monitor 24-Hour Zero and Calibration
           Drift                                                      34
                                     iv

-------
                             TABLES (continued)





Number                                                                Page





 18       Test Results for 0~ Monitor 24-Hour Zero and Calibration

           Drift            c                                         35



 19       Test Results for C09 Monitor 24-Hour Zero and Calibration

           Drift             c                                        36



 20       Monitor Response Time                                       37



 21       NO  Monitor Response Time                                   37
            A


 22       0? Monitor Response Time                                    38



 23       CO Monitor Response Time                                    38



 24       C02 Monitor Response Time                                   38



 25       NO  Linear Regression Data                                  44
            A


 26       02 Linear Regression Data                                   45



 27       CO Linear Regression Data                                   46



 28       CO^ Linear Regression Data                                  47



 29       Summary of NO  CEM Audit Results                            48
                       A


 30       Comparison of Reference Method 7 and NO  CEM Test  Results    49
                                                 A


 31       Comparison of Oxygen and Carbon Dioxide Results—CEM and

           Reference Method 3 (Orsat)                                 50



 32       Field Equipment Calibration                                 52



 33       Particulate Filter and Reagent Blank Analysis               58



 34       Boiler Process Data                                         77

-------
                               ACKNOWLEDGMENT

     Mr. Dennis Holzschuh, EPA Task Manager, provided overall  project coor-
dination and guidance and observed the test program.   Messrs.  Kevin Johnson
and John Martinez of Radian Corporation, an EPA Contractor,  provided project
coordination relative to project scope and process operation.   Mr.  Len Keck
of General Electric Company provided assistance in scheduling  and process
operation.  Mr. Charles Bruffey was the PEI Project Manager.   Principal
report authors were Messrs. Charles Bruffey, Paul  Reinermann,  and Daniel
Scheffel.

-------
                      QUALITY ASSURANCE ELEMENT FINDER
                                                                 Location
                                                             Section     Page
(1)  Title page

(2)  Table of contents

(3)  Project description

(4)  QA objective for measurement of data  in  terms
     of precision, accuracy,  completeness,  repre-
     sentativeness, and comparability

(5)  Sampling procedures


(6)  Sample custody

(7)  Calibration procedures and frequency


(8)  Analytical  procedures


(9)  Data reduction, validation, and reporting


(10)  Internal quality control  checks and frequency


(11)  Performance and system audits and  frequency


(12)  Preventive maintenance procedures  and  schedules

(13)  Specific routine procedures used to assess data
     precision, accuracy,  and  completeness  of specif-
     ic measurement parameters involved

(14)  Corrective action

(15)  Quality assurance reports to management

1
Appendix F
Section 3
Appendix D
Section 4
Appendix C
Appendix E
Section 3
Appendix D
Section 4
Appendix F
Section 3
Appendix F
Section 3
Appendix F
Section 3
Appendix F
11
1
F-2
D-l
C-l
E-l
D-l
F-3
F-5
F-3
F-12
Appendix F   F-4

Appendix F   F-ll

Appendix F   F-12
                                     vn

-------
                                  SECTION 1
                                INTRODUCTION

     The United States Environmental  Protection Agency (EPA)  is  developing
standards of performance for industrial  boilers in accordance with  Section
111 of the Clean Air Act as amended August 1977.   The Act  requires  that  the
standards be based on the "...  best technological  system of continuous emis-
sion reduction which the Administrator of EPA determines has  been adequately
demonstrated."  Accordingly, EPA is interested in  the nitrogen oxide (NO )
                                                                        /\
control capability of Low Excess Air (LEA) burners on industrial boilers.
     To support the standards development process  and provide data  to charact-
erize emissions from industrial boilers, PEI performed a series  of  atmospheric
emission tests on two oil- and gas-fired boilers equipped  with the  latest
commercial LEA burners at the General Electric Noryl  Products facility in
Selkirk, New York.  These tests were conducted under  contract to EPA's  Emis-
sion Measurement Branch (EMB) from August 14 to 20, 1984.   The primary objec-
tives of the test program were:
     0    To characterize NO  emissions as a function of fuel type, boiler
          load, excess air [oxygen (02) level], and combustion air  preheat
          temperature.
     0    To obtain particulate emissions data.
No major problems were encountered during the test program and project objec-
tives were met.
     All testing was performed on Boilers 4 and 5.  Both boilers were tested
while firing natural gas, and Boiler 5 only was tested while firing No.  6
residual and No. 2 distillate oil.  Continuous emission monitor  (CEM) systems
for NO  , 02, CO, and C02 were used to characterize these pollutants as a
function of boiler load, excess air, and combustion air reheat temperature.
Manual testing was conducted concurrently with the CEM tests while  Boiler 5

-------
was firing residual and distillate oil to determine the concentration and
mass emission rate of participate matter according to procedures described in
EPA Test Methods 5 and 5B.*  Flue gas volumetric flow rates, temperature, and
moisture content were determined in conjunction with the particulate tests.
As a data quality assurance check for the NOV CEM system, EPA Method 7 (NOV)
                                            /^                             /\
tests were also conducted over the course of the test program.  In addition,
fuel oil samples were collected during the test program and subjected to a
proximate and ultimate analysis.
     Section 2 of this report summarizes the results of the test program.
Section 3 addresses quality assurance activities undertaken to assure repre-
sentative data collection.  Section 4 summarizes the testing and analytical
procedures used and describes the sampling locations.  Section 5 describes
the process and its operation during the test series.  Appendices A through  F
contain computer printouts and calculations, all field and laboratory data
sheets, detailed descriptions of the testing and analytical procedures used,
and equipment calibration procedures and results.
 40 CFR 60, Appendix A, July 1984.

-------
                                   SECTION 2
                            SUMMARY OF TEST RESULTS

     This section details the results of the field test program.   For the
convenience of the reader, emission data are presented in both  metric and
English units where applicable.   Also, subsections are used to  present each
phase of the sampling program and the corresponding emission results.

2.1  TEST PROTOCOL
     Table 1 presents a summary of the process operating parameters and the
types of tests performed during the program.  For reporting purposes, the
sampling dates, test times, and types of tests are grouped by fuel  type (oil
or natural gas).  The actual sequence of events varied somewhat because of
boiler operating problems and plant production schedules that affected boiler
load.
     Two identical boilers (Boilers 4 and 5) capable of firing  both natural
gas and fuel oils were used for this study.  Each has a rated capacity of
150,000 pounds of steam per hour, but normal operation is in the 50,000 to
60,000 Ib/h range.  Both boilers possess Coen parallel-flow, low-excess air
burners.  Flue gas exits the top of the boiler, turns, and then passes
downward through an economizer before it enters the exit stack  via a
rectangular breeching.  Testing was conducted in the exit stacks of the two
boilers.  Boiler 4 was fired with natural gas, and Boiler 5 was fired with
both No. 6 residual oil and No.  2 distillate oil as well as natural gas.  A
total of 19 individual test blocks were conducted during which  NO , 0?, CO,
                                                                 /\   C~
and C02 emissions were continuously monitored.  Of this total,  six test blocks
were conducted on Boiler 4 and the remainder on Boiler 5.  Ten  of the test
blocks were for natural gas (Boilers 4 and 5), and nine were for residual  and
distillate oil (Boiler 5 only).
     As shown in Table 1, particulate emission tests on Boiler 5 were con-
ducted according to EPA Methods 5* and 5B* sampling and analytical  procedures

-------
TABLE 1.   SUMMARY OF BOILER  OPERATING  PARAMETERS  AND  EMISSIONS  TESTS
Test
block
1
2
3
4
5
6
7
8
9
NG-1
NG-2
Date
(1984) and
Time (24-h)
8/16
1215-1415
8/16
1707-1907
8/17
0027-0227
8/17
1112-1442
8/17
1645-1910
8/17
2201-2341
8/18
1725-1925
8/18
2049-2249
8/19
0014-0214
8/15
1310-1740
8/15
1823-2018
Boiler
ID
No. 5
No. 5
No. 5
No. 5
No. 5
No. 5
No. 5
No. 5
No. 5
No. 4
No. 4
Operating parameters
Fuel9 type
R.O.
R.O.
R.O.
D.O.
D.O.
Low viscosity
R.O.
R.O.
R.O.
R.O.
N.G.
N.G.
Loadb
1/2
1/2
1/2
1/2
1/2
1/2
Full
3/4
3/4
3/4
1/2
02 Level c
High
Low
Low
High
Low
Low
Low
Low
High
Low
Low
Air
preheat
tempera-
ture, °F
Ambient
Ambient
si 60
si 50
si 50
Ambient
Ambient
Ambient
Ambient
Ambient
Ambient
Emission tests
CEM's
(NO , 0,,
C02, CO)
X
X
X
X
X
X
X
X
X
X
X
Particulate
(Methods 5-5B)
X
X
X
X
X
X
X
X
X
-
-
Reference
Method 7
(NOX)
X
X
-
-
-
X
X
X
-
X
-
(continued)

-------
TABLE 1  (continued)
Test
block
NG-3
NG-4
NG-5
NG-6
NG-7
NG-8
NGr9
NG-10
Date
(1984) and
Time (24-h)
8/19
1041-1241
8/19
1440-1540
8/19
1626-1726
8/19
1837-1937
8/19
2052-2152
8/19
2227-2257
8/19
2315-2345
8/19-20
2351-0021
Boiler
ID
No. 4
No. 4
No. 4
No. 4
No. 5
No. 5
No. 5
No. 5
Operating parameters
Fuel9 type
N.G.
N.G.
N.G.
N.G.
N.G.
N.G.
N.G.
N.G.
Loadb
3/4
Full
Full
1/2
1/2
3/4
3/4
3/4
02 Level0
High
High
Low
High
High
High
Low
Low
Air
preheat
tempera-
ture,0 °F
Ambient
Ambient
Ambient
Ambient
Ambient
Ambient
Ambient
sfl60
Emission tests
CEM's
(NO. 0?,
C02, CO)
X
X
X
X
X
X
X
X
Particulate
(Methods 5-5B)
-
-
-
-
-
-
-

Reference
Method 7
(NOX)
—
-
-
X
-
-
-

aFuel type:  R.O. = No. 6 Residual oil; D.O. = No. 2 Distillate oil; N.G. = Natural  gas
bBoiler load:  1/2 (=70,000 Ib/h steam); 3/4 (=105,000 Ib/h steam); Full (=140,000 Ib/h steam)
C0xygen level  for natural gas^and residual oil tests:  Low (=0.5-1.5%), High (>3.0%); for distillate oil test:
  Low  (2.0%), High (>3.0%).
  Ambient air temperature  is approximately 70°-90°F.

-------
while the boiler was firing residual  and distillate oil.   No participate
measurements were made during any of the natural  gas test blocks.   Manual
emission tests for NO  (Method 7*) were conducted during  7 of the  19  test
                     /\
blocks, as a quality assurance check of the NO  CEM system.
                                              A
     Throughout the test program, boiler load and excess  air (02)  levels were
varied to characterize NO  emissions as a function of these operational param-
                         /\
eters.  Personnel from Radian Corporation, the EPA's NSPS Contractor, coordi-
nated changes in process operation with the boiler operators and monitored all
pertinent data during each test block.   As indicated in Table 1, boiler loads
of one-half capacity (representing =70,000 Ib/h), three-fourths  capacity
(representing =105,000 Ib/h), and full  capacity (representing =140,000 Ib/h)
were evaluated.
     Excess air (0^) levels were designated as high (ranging from  3 to 4
percent) and low (typically less than 1.5 percent).  Combustion  air temperatures
were generally at ambient levels between 21° and 32°C (70°-90°F);  however, 2
of the 19 test blocks were conducted at a preheat temperature of 71°C (160°F).
     The following subsections present the results of the test program.

2.2  CONTINUOUS EMISSION MONITOR DATA
     An extractive monitoring system was assembled on site to characterize
emissions data for NO , 02, CO, and C02 as a function of  boiler load, excess
air (0?) level, and combustion air preheat temperature.  Table 2 summarizes
the boiler operating parameters and corresponding CEM data for each designated
test block.  Table 3 presents the maximum and minimum pollutant concentrations
recorded for each test block.
     A test block typically represented a 2-hour monitoring period during
which the boiler load and oxygen level  remained relatively constant at the
specified conditions listed in Table 2.  Several  of the test blocks run on
Boilers 4 and 5 while these boilers were firing natural gas were 30 minutes to
1 hour in duration.
     Regardless of the test time, the reduction of the CEM data was accom-
plished by taking an average chart reading for every 5-minute period  and
determining the corresponding pollutant concentration by  the use of linear
 40 CFR 60, Appendix A, Reference Methods 5, 5B, and 7, July 1984.

-------
                             TABLE 2.   SUMMARY OF CONTINUOUS  EMISSION  MONITORING  DATA
Test
block
1
2
3
4
1 5
6
7
8
9
NG-1
NG-2
Date
(1984) and
Time (24-h)
8/16
1215-1415
8/16
1707-1907
8/17
0027-0227
8/17
1112-1442
8/17
1645-1910
8/17
2201-2341
8/18
1725-1925
8/18
2049-2249
8/19
0014-0214
8/15
1310-1740
8/15
1823-2018
Boiler
ID
No. 5
No. 5
No. 5
No. 5
No. 5
No. 5
No. 5
No. 5
No. 5
No. 4
No. 4
Operating parameters
Fuel3 type
R.O.
R.O.
R.O.
D.O.
D.O.
Low viscosity
R.O.
R.O.
R.O.
R.O.
N.G.
N.G.
Loadb
1/2
1/2
1/2
1/2
1/2
1/2
Full
3/4
3/4
3/4
1/2
02 Level0
High
Low
Low
High
Low
Low
Low
Low
High
Low
Low
Air
preheat
tempera-
ture, °F
Ambient
Ambient
=160
=150
=150
Ambient
Ambient
Ambient
Ambient
Ambient
Ambient
Average emission data, dry basis
NO ,
ppm
354
376
385
119
130
420
429
400
380
196
169
N0y
at 3%
02, ppm
362
338
346
123
124
382
388
367
400
175
151
02, %
3.4
1.0
1.0
3.6
2.1
1.2
1.1
1.4
3.9
0.9
0.8
C02, %
13.3
15.0
15.3
12.9
13.7
15.6
14.7
15.4
13.1
11.4
11.6
CO, ppm
0
26
56
1
170
20
46
87
0
25
7
(continued)

-------
     TABLE  2  (continued)
Test
block
NG-3
NG-4 -
NG-5
NG-6
NG-7
NG-8
NG-9
NG-10
Date
(1984) and
Time (24-h)
8/19
1041-1241
8/19
1440-1540
8/19
1626-1726
8/19
1837-1937
8/19
2052-2152
8/19
2227-2257
8/19
2315-2345
8/19-20
2351-0021
Boiler
ID
No. 4
No. 4
No. 4
No. 4
No. 5
No. 5
No. 5
No. 5
Operating parameters
Fuel9 type
N.G.
N.G.
N.G.
N.G.
N.G.
N.G.
N.G.
N.G.
Loadb
3/4
Full
Full
1/2
1/2
3/4
3/4
3/4
02 Level0
High
High
Low
High
High
High
Low
Low
Air
preheat
tempera-
ture, °F
Ambient
Ambient
Ambient
Ambient
Ambient
Ambient
Ambient
S160
Average emission data,6 dry basis
NO ,
ppm
151
146
211
122
100
118
176
209
N0y
at 3%
02, ppm
152
147
188
126
104
121
159
189
02, %
3.1
3.1
0.8
3.6
3.7
3.4
1.1
1.1
C02, %
10.0
10.1
11.6
9.9
10.0
10.5
11.7
11.5
CO, ppm
0
0
12
0
0
0
19
25
00
      aFuel type:  R.O. = No. 6 Residual oil; D.O.  = No.  2 Distillate oil;  N.G.  = Natural  gas
      bBoiler load:  1/2 (=70,000 Ib/h steam); 3/4  (=105,000 Ib/h steam);  Full  (=140,000 Ib/h steam)
      ^Oxygen level for natural gas and residual oil tests:  Low (S0.5-1.5%), High (>3.0%); for distillate oil tests:
       Low (-2%), High (>3.0%).
       Ambient air temperature, between 70°-90°F.
      eAverage emission results for the indicated test block time interval, dry basis.

-------
regression equations established from the calibration  data  for  each monitor.

For a 2-hour test block, this procedure yielded 24  data  points;  for a  1-hour

test block, 12 data points; and for a 30-minute test block,  6 data points.

Pollutant concentrations reported in Table 2 represent average  values  deter-

mined from the number of data points for each test  block.   The  concentrations

reported in Table 3 represent the maximum and minimum  values recorded  during

the designated test periods.

     Nitrogen oxide (NO ) concentrations are reported  in parts  per million  by
                       A
volume on a dry basis.  These concentrations have also been corrected  to  3

percent CL as a standard by which emission trends can  be evaluated.  Oxygen

and carbon dioxide (CO-) concentrations are reported in  percent by volume and

carbon monoxide (CO) concentrations are reported in parts per million  by

volume, all on a dry basis.

     In the evaluation of the data relative to the  specific process param-

eters, NO  concentrations corrected to 3 percent 09 were used.   Analysis  of
         A                                       £.
the test data showed several general trends:

     1.   Concentrations of NO  were significantly  higher at the same  load
          conditions when Boiler 5 was burning No.  6 residual oil  rather  than
          No. 2 distillate oil.  When corrected to  3 percent 02, NO  concentra-
          tions ranged from 338 to 362 ppm for the  three residual  oil  tests
          compared with values of 123 and 124 ppm for  the two distillate  oil
          tests at half load.

     2.   Combustion air preheat temperature appeared  to have no significant
          impact on NO  concentration during residual  oil  firing.   For similar
          boiler load and oxygen levels, the results from Test  Blocks  2 and 3
          showed an average NO  concentration of 338 ppm at ambient preheat
          temperature compared with 346 ppm at a preheat temperature of 71°C
          (160°F), or less than a 2.5 percent difference.  For  natural gas,
          however, the results of Test Blocks NG-9  and NG-10 indicate  that  NOX
          concentrations increase by more than 10 percent with  an increase  in
          preheat temperature.  The average concentration at a  preheat tempera-
          ture of 71°C (160°F) was 189 ppm compared with 159 ppm at an ambient
          preheat temperature, or a 16 percent difference.

     3.   When Boiler 5 was firing residual oil, NO concentrations were
          generally greater at high excess-air leveTs.  Under similar  boiler
          load conditions, the results from Test Blocks  1 (one-half load, high
          excess air) and 2 (one-half load, low excess air) were 362 and  338
          ppm, respectively.  For Test Blocks 8 (three-fourths  load, low 02)
          and 9 (three-fourths load, high 02), NO  concentrations were 367  and
          400 ppm, respectively.  For Test Blocks 4 and  5 (while Boiler 5 was
          firing distillate oil), NO  concentrations were 123  ppm for  Test
          Block 4 (one-half load, hitjh 02) and 124  ppm for Test Block  5 (one-
          half load, low 02).

-------
     TABLE 3.    SUMMARY  OF  MAXIMUM AND MINIMUM CEM DATA BY TEST BLOCK


Test Block
RO-1
RO-2
RO-3
DO-4
DO-5
RO-6
RO-7
RO-8
RO-9
NG-1
NG-2
NG-3
NG-4
NG-5
NG-6
NG-7
NG-8
NG-9
NG-10

Date
(1984)
8/16
8/16
8/17
8/17
8/17
8/17
8/18
8/18
8/19
8/15
8/15
8/19
8/19
8/19
8/19
8/19
8/19
8/19
8/19
CEM Data3
NO , ppm
Max.
370
381
391
126
135
424
435
414
388
200
172
153
147
213
123
100
119
178
210
Min.
338
366
377
116
124
418
420
389
374
191
157
149
145
210
120
99
116
175
208
02, %
Max.
3.7
1.1
1.4
4.0
2.5
1.4
1.2
1.5
3.9
1.2
0.9
3.1
3.1
0.9
3.6
3.7
3.4
1.2
1.2
Min.
3.3
0.9
0.9
3.1
1.9
1.0
1.0
1.2
3.9
0.8
0.8
3.1
3.1
0.8
3.6
3.7
3.3
0.9
0.8
C02, %
Max.
13.7
15.2
16.1
13.1
14.0
15.8
14.9
15.6
13.2
11.7
12.0
10.4
10.2
11.8
10.1
10.0
10.5
11.8
11.7
Min.
12.7
14.4
14.4
12.2
13.4
15.4
14.4
15.3
13.0
11.1
11.0
9.7
9.9
11.1
9.6
9.9
10.5
11.7
11.3
CO, ppm
Max.
0
58
134
0
234
133
87
124
0
68
9
0
0
13
0
0
0
23
47
Min.
0
10
5
0
35
2
27
20
0
3
6
0
0
11
0
0
0
12
13
Dry basis.
                                  10 .

-------
     4.    When firing natural  gas,  NO   concentrations were generally higher at
          low excess-air levels.  For Test Blocks NG-4  (full load, high 02)
          and NG-5 (full  load, low  02),  NO   concentrations were 147 and 188
          ppm, respectively.   A similar pattern was observed for Test Blocks
          NG-8 (three-fourths  load, high 0?)  and NG-9 (three-fourths load, low
          02)9 during which NO  concentrations were 121 and 159 ppm, respec-
          tively.

     5.    When Boiler 5 was firing  residual  oil, the increase in NO  concen-
          tration  was less than 10  percent between one-half boiler Toad and
          three-fourths and full  boiler loads.  Test Blocks 1, 2, 3, and 6
          (one-half loads) showed an average NO  concentration of 357 ppm.
          Test Blocks 7, 8, and 9 (three-fourths and full boiler loads) showed
          an average NO  concentration  of 385 ppm.  No  full or three-fourths
          load test data are available  for distillate oil; therefore, compari-
          sons based on boiler loads are not possible.

     6.    When natural gas was being fired,  NO  concentrations averaged 127
          ppm for  one-half boiler loads (Tests NG-2, 6, and 7), 159 ppm for
          three-fourths boiler loads (Tests  NG-1, 3, 8, 9, and 10), and 168
          ppm for  full boiler loads (Tests NG-4 and 5).

     Additional data reduction for  NO   and 09 was performed for those periods
                                    A       £.
prior to actual test blocks when the boiler  was in a changeover mode, either
from high 0^ to low 0^ or vice versa.   The purpose of this was to establish a

clearer relationship between NO  concentrations, excess-air levels, and fuel
                               /\
types.  Tables 4 and 5 summarize these  data.
          TABLE 4.  RELATIONSHIP BETWEEN NOY  CONCENTRATION  AND  EXCESS
                            AIR LEVELS—RESlDUAL  OIL
Test Block
Prior to Test 2
(one-half load,
R.O. , low 02)
Date
(1984)
8/16
8/16
8/16
8/16
8/16
8/16
Time
(24-h)
1615
1625
1638
1645
1655
1700
Average CEM concentrations
02, %
3.5
3.4
1.2
0.8
1.5
0.9
NO , ppm
As measured
340
340
360
374
337
366
At 3% 02
350
348
327
333
311
328
     Only one set of data was available for residual  oil  firing,  but these

data are generally comparable to the actual test block data,  which  show a

decrease in NO  emissions at low excess-air levels.   Seven sets  of  data were
              A
                                    11

-------
          TABLE 5.   RELATIONSHIP  BETWEEN  NOY  CONCENTRATION  AND  EXCESS
                            AIR  LEVELS—NATURAL  GAS
Test Block
Prior to Test NG-1
(three-fourths
load, N.G. , low
02)



NG-1 (Process upset
from 13:35-16:15,
boiler parameters
readjusted)




Prior to Test NG-3
(three-fourths
load, N.G. , high
02)




Prior to Test NG-5
(full load, N.G.,
high 02)

Date
(1984)
8/15
8/15
8/15
8/15
8/15
8/15
8/15
8/15
8/15
8/15
8/15
8/15
8/15
8/15
8/15
8/19
8/19
8/19
8/19
8/19
8/19
8/19
8/19
8/19
8/19
8/19
8/19
Time
(24-h)
1230
1240
1250
1300
1310
1320
1330
1505
1515
1525
1535
1545
1555
1605
1615
0935
0945
0955
1005
1015
1025
1035
1041
1606
1610
1616
1626
Average CEM concentrations
02, %
3.2
2.9
2.6
1.8
1.1
1.2
1.2
3.7
3.1
2.8
2.4
1.9
1.7
1.6
0.9
1.1
1.4
1.8
2.1
2.2
2.6
2.9
3.0
3.2
2.7
1.8
0.9
NO , ppm
As measured
144
151
164
181
193
192
194
136
156
167
177
185
185
186
196
188
183
178
178
173
166
153
149
145
166
195
212
At 3% 02
147
150
160
170
174
174
176
142
157
165
171
174
172
172
175
170
168
167
169
166
162
152
149
147
163
183
190
(continued)
                                      12

-------
TABLE 5 (continued)
Test Block
Prior to Test NG-6
(three-fourths
load, N.G. , high
02)


Prior to Test NG-7
(one-half load,
N.G., high 02)







During to Test NG-10
(three-fourths
load, N.G. , low
02)
Date
(1984)
8/19
8/19
8/19
8/19
8/19
8/19
8/19
8/19
8/19
8/19
8/19
8/19
8/19
8/19
8/19
8/19
8/19
8/19
8/19
8/19
Time
(24-h)
1821
1823
1825
1827
1831
1835
2020
2024
2026
2028
2035
2038
2040
2042
2046
2049
2345
2351
0001
0011
Average CEM concentrations
02, %
1.6
2.4
2.6
3.0
3.8
3.5
2.2
3.5
4.0
4.3
4.6
4.0
3.5
3.2
3.4
4.0
0.4
0.6
0.9
1.2
NO , ppm
As measured
146
143
142
140
117
122
127
105
95
90
83
93
101
110
105
92
199
210
210
208
At 3% 02
135
138
139
140
122
126
122
108
101
97
91
98
104
111
107
97
174
185 .
188
189
                                     13

-------
available for the natural gas tests, and these data correlate with the actual
test block data, which show an increase in NO  emissions at low excess-air
                                             A
levels.
     For data presention and informational purposes, the NO  CEM data  were
                                                           X
used to calculate mass emission rates in pounds per million Btu.   This was
accomplished by converting the reported average parts per million of NO
                                                                       /\
concentration (corrected to 3 percent 02) to pounds per dry standard cubic
foot and then multiplying this value by the appropriate F-factor based on  fuel
type and an excess air correction factor.  The F-factor relates the amount of
dry flue gas generated to the calorific value of the fuel combusted and is
expressed in dry standard cubic feet per million Btu heat input.
     Subpart D of the Federal Register* lists an F-factor of 9220 dscf/106 Btu
for residual and distillate oil and an F-factor of 8740 dscf/10  Btu for
natural gas.  These factors were used in all calculations.  Table 6 summarizes
the NOV mass emission rate data for each test block.  For the distillate oil
                                         6
tests, NO  emission rates were 0.16 lb/10  Btu.  For the residual oil  tests,
         x                                       /-
NO  emission rates ranged from 0.43 to 0.51 lb/10  Btu and emission rates
  x                           c
ranged from 0.13 to 0.23 lb/10  Btu for the natural gas tests.

2.3  PARTICULATE AND FUEL ANALYSIS TEST RESULTS
     Particulate emission measurements were made on Boiler 5 while it was
firing residual and distillate oil.  Two individual sampling trains were used
for simultaneous traversing of the cross-sectional area of the stack during
each of the nine test blocks.  One sampling train was an EPA Reference Method
5** [sample collection temperature of 121°C (250°F)] and the other was an  EPA
Reference Method 5B** [sample collection temperature of 160°C (320°F)].  The
analytical procedures for these methods are similar in all respects except
that the probe rinse and filter particulate catch for Method 5B are heated in
an oven for 6 hours at 160°C prior to gravimetric weighing.
     Previous EPA studies at sources where the emission streams contained
sulfur oxides indicate significant positive biases can occur on particulate
  40 CFR 60, Subpart D, New Source Performance Standards for Fossil-Fuel-
**Fired Steam Generators, July 1984.
  40 CFR 60, Appendix A, Reference Methods 5 and 5B, J.uly 1984.
                                     14

-------
                      TABLE 6.   SUMMARY OF NOV EMISSION RATES
                                            X
Test Block
RO-1
RO-2
RO-3
DO-4
DO-5
RO-6
RO-7
RO-8
RO-9
NG-1
NG-2
NG-3
NG-4
NG-5
NG-6
NG-7
NG-8
NG-9
NG-10
Date
(1984)
8/16
8/16
8/17
8/17
8/17
8/17
8/18
8/18
8/19
8/15
8/15
8/19
8/19
8/19
8/19
8/19
8/19
8/19
8/19
NOV concentration3
ppm at 3% 02
362
338
346
123
124
382
388
367
400
175
151
152
147
188
126
104
121
159
189
Ib/dscf x 10"b
4.32
4.04
4.13
1.47
1.48
4.56
4.62
4.37
4.78
2.09
1.80
1.82
1.76
2.25
1.51
1.24
1.45
1.90
2.26
NO mass .
emission rate,
lb/106 Btu
0.47
0.43
0.44
0.16
0.16
0.49
0.50
0,47
0.51
0.21
0.18
0.19
0.18
0.23
0.15
0.13
0.15
0.19
0.23
 Conversion factor:
                          M
                     385.1 x 10"
where M = the molecular weight of N02(46).


bF-factor:  9220 dscf/10^ Btu for R.O. and D.O.

            8740 dscf/10° Btu for N.G.
       Ib NO /106 Btu = N0¥ cone.  (Ib/dscf @ 3% 09)
            /\             /\                      £
                       x F-factor (dscf/106 Btu @ Stoichiometric 0~)
                           20.9
	   /Stoichiometric 02	\

20.9-3.0   ^dilution by excess air
                                      15

-------
measurements as a result of the retention of condensible sulfate material in
the front half (probe and filter) of the standard Method 5 sampling train.
In most cases, sulfuric acid (f-LSO.) is the predominant sulfate species in
these emission streams.  These same EPA studies have shown that increased
sampling temperatures coupled with a thermogravimetric analytical  procedure
reduce, but do not eliminate, the effects (artifact formation) of H,,S04 col-
lected in the sampling train.  Therefore, conducting both test methods makes a
direct comparison between methods possible with regard to potential biases
caused by condensible sulfate material.
     Tables 7 and 8 summarize the particulate sampling conditions and emission
results, respectively.  Sample volumes are reported in dry normal  cubic meters
(dNm3).  Volumetric flow rates are presented in actual cubic meters per minute
(m3/min), actual  cubic feet per minute (acfm), dry normal cubic meters per
minute (dNm3/min), and dry standard cubic feet per minute (dscfm).
     Particulate weights are reported in milligrams, concentrations are re-
ported in milligrams per dry normal cubic, and grains per dry standard cubic
feet.  Mass emission rates are reported in kilograms per hour, pounds per
hour, and pounds per million Btu heat input.  The mass emission rates in
kilograms/hour (pounds/hour) were calculated by multiplying individual par-
ticulate concentrations by the average flow rate measured during each specific
test block.  Mass emission rates in pounds/10  Btu were determined  by using
the particulate concentration, the 0~ level measured by the CEM system for the
specific test block, and the listed F-factor for residual and distillate oil,
9220 dscf/106 Btu.*
     As reported in Table 7, sample volumes were generally consistent for each
train, and isokinetic sampling rates ranged between 92.1 and 104.7  percent,
which is within the acceptance range of 90 to 110 percent.
     The probe and filter temperatures represent average values determined
from data recorded on the field data sheets.  Method 5 filter temperatures
ranged from 121° to 141°C (250° to 284°F), and probe temperatures ranged
from 115° to 154°C (238° to 301°F), Method 5B filter temperatures ranged from
157° to 176°C (314° to 349°F), and probe temperatures ranged from 158° to
181°C (316° to 358°F).
*40 CFR 60, Subpart D, July 1984.
                                      16

-------
                    TABLE 7.   SUMMARY  OF  PARTICULATE  SAMPLING  CONDITIONS,  BOILER  NO.   5 STACK
Run No.
M5-1
M5B-1
MS -2
M5B-2
M5-3
MSB-3
MS -4
MSB -4
M5-5
HSB-S
M5-6
MSB-6
M5-7
M5B-7
M5-8
M5B-8
M5-9
M5B-9
Sample
type
MS
MSB
MS
MSB
MS
MSB
MS
MSB
MS
MSB
M5C
MSB
MS
MSB
MS
MSB
MS
MSB
Date
(1984)
8/16
8/16
8/16
8/16
8/16
8/16
8/17
8/17
8/17
8/17
8/17
8/18
8/18
8/18
8/18
8/19
8/19
Time
(24-h)
1219-1555
1217-1553
1732-2235
1733-2236
0035-0310
0036-0311
1128-1435
1129-1436
1649-1926
1650-1925
2210-0035
1728-1902
1729-1901
2050-2212
2051-2213
0025-0150
0026-0151
Samplea
volume,
dNm3
3.34
3.19
1.78
1.71
1.60
1.68
1.90
1.92
1.65
1.65
1.61
1.71
1.71
1.33
1.43
1.58
1.63
Sampl ing
time,
min
120
120
120
120
120
120
120
120
120
120
120
72
72
72
72
72
72
Isokinetic
sampl ing
rate, %
99.5
92.1
104.7
104.3
101.7
99.5
101.5
103.2
104.3
103.4
101.9
98.4
97.9
99.0
98.5
99.0
100.1
Sampling temperature
Probe
"C (°F)
130 (266)
172 (342)
121 (250)
158 (316)
120 (249)
163 (326)
115 (238)
176 (350)
160 (320)
168 (335)
161 (322)
154 (310)
181 (358)
149 (301)
164 (327)
137 (278)
162 (323)
Filter
°C( °F)
123 (253)
172 (341)
125 (257)
164 (328)
127 (261)
157 (314)
135 (275)
167 (333)
121 (250)
163 (325)
169 (336)
140 (284)
176 (349)
130 (266)
169 (337)
128 (263)
167 (332)
b
Flue gas conditions
Volumetri flow rate
mVinin (acfm)
971 (34,300)
869 (31.100)
833 (29,400)
960 (33.900)
823 (29,100)
810 (28,600)
1704 (60,200)
1281 (45,250)
1524 (53.800)
dNmVmin (dscfm)
579 (20.400)
544 (18,000)
492 (17.400)
572 (20,200)
486 (17,200)
481 (17.000)
889 (31,400)
713 (25,200)
823 (29.100)
Temperature,
°C (°F)
157 (315)
157 (314)
149 (300)
157 (313)
153 (307)
147 (296)
214 (417)
182 (361)
201 (393)
Moisture
content, %
12.3
14.5
14.7
12.3
13.6
14.3
12.7
12.8
12.0
aSample volume in dry normal  cubic meters per minute.   Standard conditions:  760 mraHg (29.92 in.Hg); 20°C  (68°F); and 0 percent moisture.
bVolumetr1c  flow rate, gas temperature, and moisture content data represent the average  values measured by each Individual sampling train.
cRun MS-6 exhibited an excessive leak  rate at the sample port change, and the run is considered void.

-------
TABLE 8.  SUMMARY OF PARTICULATE EMISSION RESULTS
Test
block
1
2
3
4
5
6
7
8
9
Date
(1984)
8/16
8/16
8/16
8/17
8/17
8/17
8/18
8/18
9/19
Sample
ID
M5-1
M5B-1
M5-2
M5B-2
M5-3
M5B-3
M5-4
M5B-4
M5-5
M5B-5
M5-6a
M5B-6
M5-7
M5B-7
M5-8
M5B-8
M5-9
M5B-9
Analytical results
Particulate
weight, mg
Probe
30.7
9.1
18.2
4.0
16.9
6.0
13.8
6.3
15.8
6.0
6.8
58.0
40.3
34.8
13.6
32.0
12.1
Filter
228.9
169.6
127.7
103.2
100.9
97.8
9.4
13.6
4.1
4.0
98.4
282.7
211.9
111.3
100.4
118.0
101.1
Total
258.9
178.7
145.9
107.2
117.8
103.8
23.2
19.9
19.9
10.0
105.2
340.7
252.2
146.1
114.0
150.0
113.2
Concentration
mg/dNm3
77.5
56.1
82.1
62.8
75.8
62.0
12.2
10.4
12.1
6.1
65.5
198.9
148.0
109.7
79.9
94.9
69.5
gr/dscf
0.034
0.0245
0.036
0.027
0.033
0.027
0.005
0.0045
0.005
0.003
0.029
0.087
0.065
0.048
0.035
0.0415
0.030
Mass emission rate
kg/h
2.7
2.0
2.6
1.9
2.1
1.9
0.42
0.35
0.35
0.18
1.7
10.6
7.9
4.5
3.5
4.7
3.5
Ib/h
5.9
4.3
5.7
4.2
4.7
4.2
0.93
0.78
0.78
0.39
4.2
23.5
17.3
10.0
7.8
10.3
7.6
lb/10b Btu
0.05
0.04
0.05
0.04
0.05
0.04
0.01
0.01
0.01
0.01
0.04
0.12
0.07
0.07
0.05
0.07
0.05

-------
     The reported flue gas volumetric flow rates,  temperatures,  and  moisture
contents represent the average of the individual  train measurements.   In  each
case, individual train measurements compared within ±10 percent.
     Test Blocks 1, 2, 3, and 6 were conducted at  approximately  one-half
boiler load during the firing of residual  oil.  Average flue gas flow rates
ranged from 810 m2/min (28,600 acfm) during Test 6 to 971 m3/min (34,300  acfm)
during Test 1. ' Temperatures ranged from 147° to 157°C (296° to  315°F), and
moisture content ranged from 12.3 to 14.7  percent.  Particulate  concentrations
and mass emission rates as measured by the Method  5 sampling trains  for these
test blocks ranged from 75.8 mg/dNm3 (0.033 gr/dscf) and 2.1 kg/h (4.7 lb/h;
0.05 lb/106 Btu) to 82.1 mg/dNm3 (0.036 gr/dscf) and 2.6 kg/h (5.7 Ib/h;  0.05
lb/10  Btu).  The corresponding Method 5B  test results showed particulate
concentrations and mass emission rates ranging from 56.1 mg/dNm3 (0.0245
gr/dscf) and 2.0 kg/h (4.3 lb/h; 0.04 lb/106 Btu)  to 65.5 mg/dNm3 (0.029
gr/dscf) and 1.7 kg/h (4.2 lb/h; 0.04 lb/106 Btu).  Run M5-6 was void because
of an excessive mid-test leakage rate; therefore,  results are not reported.
     Test Blocks 8 and 9 were conducted at approximately three-fourths boiler
load while Boiler 5 was firing residual oil.  Average flue gas flow rates
ranged from 1281 m3/min (45,250 acfm) during Test 8 to 1524 m3/min (53,800
acfm) during Test 9.  Gas temperature and  moisture content were  182°C (361°F)
and 12.8 percent during Test 8 and 201°C (393°F) and 12.0 percent during  Test
9.  Method 5 results during Test 8 showed  a particulate concentration of  109.7
mg/dNm3 (0.048 gr/dscf) and a mass emission rate of 4.5 kg/h (10.0 lb/h;  0.07
lb/106 Btu).  The corresponding Method 5B  results  were 79.9 mg/dNm3  (0.035
gr/dscf) and 3.5 kg/h (7.8 Ib/h; 0.05 lb/106 Btu).  During Test  9, Method 5
showed a concentration of 94.9 mg/dNm3 (0.0415 gr/dscf) and an emission  rate
of 4.7 kg/h (10.3 lb/h; 0.07 lb/106 Btu);  Method 5B results were 69.5 mg/dNm3
(0.030 gr/dscf) and 3.5 kg/h (7.6 Ib/h; 0.05 lb/106 Btu), respectively.
     Test Block 7 was conducted at full boiler load while Boiler 5 was firing
residual oil.  The average gas flow rate,  temperature, and moisture content
during this run were 1704 m3/min (60,200 acfm), 214°C (417°F), and 12.7  per-
cent, respectively.  Method 5 results showed a particulate concentration  of
198.9 mg/dNm3 (0.087 gr/dscf) and a mass emission rate of 10.6 kg/h (23.5
lb/h; 0.12 lb/106 Btu).  The Method 5B results were 148.0 mg/dNm3 (0.065
gr/dscf) and 7.9 kg/h (17.3 lb/h; 0.07 lb/106 Btu).
                                     19

-------
     Test Blocks 4 and 5 were conducted at approximately one-half boiler load
while Boiler 5 was firing distillate oil.  For Test 4, the flue gas flow rate
averaged 960 m3/min (33,900 acfm), temperature averaged 157°C (313°F), and
moisture content averaged 12.3 percent.  Method 5 results showed a particulate
concentration of 12.2 mg/dNm3 (0.005 gr/dscf) and a mass emission rate of 0.42
kg/h (0.93 Ib/h; 0.01 lb/106 Btu).  The Method 5B results were 10.4 mg/dNm3
(0.0045 gr/dscf) and 0.35 kg/h (0.78 Ib/h; 0.01 lb/106 Btu).   For Test 5, the
gas flow rate, temperature, and moisture content averaged 823 m3/min (29,100
acfm), 153°C (307°F), and 13.6 percent, respectively.   Method 5 results showed
a particulate concentration of 12.1 mg/dNm3 (0.005 gr/dscf) and a mass emis-
sion rate of 0.35 kg/h (0.78 Ib/h; 0.01 lb/106 Btu).  The Method 5B results
were 6.1 mg/dNm3 (0.003 gr/dscf) and 0.18 kg/h (0.39 Ib/h; 0.01 lb/106 Btu).
     Table 9 presents a summary of the between-run statistical data for simi-
lar sample types and boiler loads.  For each set of data, the table lists the
mean particulate concentration, the standard deviation with N-l weighting for
sampling data, and the percentage coefficient of variance (CV), which ex-
presses the standard deviation as a percent of the mean concentration.  The
number of data points included in each calculation is  shown for consideration
in data evaluation.  The data show that regardless of  the difference in con-
centrations as determined by Methods 5 and 5B, the between-run data for simi-
lar load conditions and sample types are reproducible.
     Table 10 summarizes the fuel analytical results for residual and distil-
late oil samples collected during the particulate test program.  Samples of
the oil were collected immediately prior its entering  the boiler burner sys-
tem.  For Test Blocks 1 through 3 and 7 through 9, samples of residual oil
were collected at the indicated times, and a single composite sample was
analyzed by applicable ASTM Methods.  Individual samples were collected for
Test Blocks 4 through 6.
     The residual oil samples showed an ash content ranging from 0.052 to 0.14
percent, a sulfur content ranging from 2.18 to 2.42 percent,  and a nitrogen
content ranging from 0.34 to 0.43 percent.  The distillate oil samples showed
an ash content of less than 0.02 percent, an average sulfur content of 0.55
percent, and a nitrogen content of less than 0.02 percent.
                                     20

-------
            TABLE 9.  BETWEEN-RUN  STATISTICAL DATA FOR SIMILAR SAMPLE
                            TYPES  AND  BOILER LOADS
Run No.
M5-1
M5-2
M5-3
M5B-1
M5B-2
M5B-3
M5B-6
M5-8
M5-9
M5B-8
M5B-9
M5-4
M5-5
M5B-4
M5B-5
Boiler load
1/2
1/2
1/2
1/2
1/2
1/2
1/2
3/4 .
3/4
3/4
3/4
1/2 - Distillate oil
1/2 - Distillate oil
1/2 - Distillate oil
1/2 - Distillate oil
Participate
concentra-
tion, mg/dNm3
77.5
82.1
75.8
56.1
62.8
62.0
65.5
109.7
94.9
79.9
69.5
12.2
12.1
10.4
6.1
No. of
data
points
3
4
2
2
2
2
Statistical data
x,a
mg/dNm3
78.5
61.6
102.3
74.7
12.15
8.25
b
o»
mg/dNm3
3.26
3.96
10.5
7.35
0.07
3.04
CV,C %
4.2
6.4
10.2
9.8
0.6
37.0
 Mean participate concentration.

 Between-run standard deviation with N-l  weighting  for  sampling data.

Coefficient of variance is the standard  deviation  expressed as a percentage of
 the mean.
                                     21

-------
                                  TABLE 10.  SUMMARY OF FUEL ANALYTICAL  RESULTS

Laboratory ID
Composite No. 1



No. 2 DS064
No. 3 DS065
Mo. 4 DS063
Composite No. 5



Date
(1984)
8/16
8/16
8/16
8/17
8/17
8/17
8/17
8/18
8/18
8/19

Time
(24-h)
1130
1908
2200
0330
1430
1930
2355
1940
2250
0225

Fuel type

No. 6 residual oil


No. 2 distillate oil
No. 2 distillate oil
No. 6 residual oil

No. 6 residual oil

Fuel analysis, % by weight
Ash

0.14


0.018
0.013
0.052

0.082

Carbon

85.75


85.24
85.23
85.57

83.64

Sulfur

2.18


0.53
0.57
2.18

2.42

Hydrogen

7.08


11.97
12.84
11.43

9.92

Oxygen

4.42


2.24
1.35
0.42

3.51

Nitrogen

0.43


<0.02
<0.02
0.34

0.42

Heating
value,
Btu/lb

18,690


19,280
18,660
18,660

18,610

ro
ro

-------
     The reported analytical results and heating values generally correspond
to the range of values expected for these specific fuel types according  to
reference material.*
     Analysis of the particulate test results revealed the following readily
evident factors:
     1.   Regardless of the test type, particulate emissions increased signif-
          icantly as boiler load increased.
     2.   Particulate emissions generated by the firing of residual  oil  were
          substantially higher than those generated by the firing of distil-
          late oil.
     3.   The particulate sample results (concentration basis) produced  by  EPA
          Method 5B were consistently lower by an average of approximately  30
          percent than the within-run Method 5 sample results.
     Items 1 and 2 were as expected given previous test experience and the
fuel analysis results presented in Table 10.  With regard to Item 3, past EPA
studies (as mentioned previously) have shown the effects of condensible
sulfate biases on particulate measurements from sources with emission streams
laden with sulfur oxides.
     The sulfur data in the fuel analysis would lead one to expect sulfate
species in the gas emission stream from this boiler.  Sulfuric acid (hLSO,)
and/or its metal and ammonium salts are the most probable forms of water-
soluble sulfates that exist in many complex chemical forms.  Of these, HpSO^
is the most common.
     The EPA studies cited in the report have shown that increased sample
temperature coupled with a thermogravimetric analysis (i.e., Method 5B)  will
minimize the positive biases on particulate measurements caused by the reten-
tion of ^SO. and its associated water in the front half of the standard
Method 5 sampling train.
     These same studies have shown that the greater the amount of ^SO.  re-
tained in the probe and in the filter, the greater the positive bias in  par-
ticulate measurements.  Increased hLSO. retention also increases the potential
for artifact formation.  The data suggested that the greater the amount  of
*
 The Babcock and Wilcox Company.  Useful  Tables for Engineers.   13th ed.
 1979.  p. 58.

                                     23

-------
condensible sulfate (HpSCh) initially collected,  the more that remained  on  the

samples, regardless of treatment temperature.   The analytical  data  indicated

that HUSCL and its associated water were significantly reduced by heating

sample fractions to at least 160°C (320°F)  prior  to gravimetric analysis.

Observed weight losses at higher treatment  temperatures were primarily attrib-

utable to the volatilization of residual HUSO,  and other water-soluble sul-

fates not removed by heating at 160°C.

     According to current understanding of  the  thermogravimetric principle,

only sulfate present as HpSO^ and its associated  water is removed in  the heat

procedure.  Other water-soluble sulfates would  not be removed  at 160°C.

Therefore, if H^SO, is the predominant sulfate  species, the Method  5B results

would be expected to be lower than the Method  5 results, and correcting  for

total water-soluble sulfate should account  for  a  significant portion  of  the

differences in weights.

     In an effort to characterize this difference in test results,  the follow-

ing additional analyses were performed.

     1.   A single weighing of each Method  5B  sample fraction  was performed
          prior to heating to 160°C.  In this way, the weight  loss  was evalu-
          ated as a function of heat treatment  temperature.

     2.   A single Method 5 test (M5-9) was subjected to heat  treatment  at
          160°C to determine sample weight  loss for comparison with the  with-
          in-run Method 5B data (M5B-9).

     3.   The probe rinse and filter samples from Runs 1 through 9  were  ana-
          lyzed for residual sulfate (SOi* ).  Each filter was  cut into small
          pieces and placed in a 125-ml flask  equipped with a  stopper and  air
          condenser.  About 50 ml of deionized, distilled water was added,  and
          the contents were gently refluxed for 6 to 8 hours.   The  solution
          was then cooled and diluted with  water  to exactly 250 ml  in a  volu-
          metric flask.  A 15-ml aliquot of the settled solution was  then
          pipetted in a separate container.  The  sulfate ion (SOit") concentra-
          tion in each aliquot was then determined by ion chromatography.   The
          results of this analysis yielded  total  milligrams of S04  ,  or  total
          water-soluble sulfate, on the filter  samples.  The probe  rinse
          residue was diluted to 150 ml with deionized, distilled water  and
          the contents were heated to approximately 90°C and gently stirred
          for 2 to 3 hours.  This solution  was  cooled and diluted with deion-
          ized, distilled water to 250 ml in a  volumetric flask.  A 15-ml
          aliquot of the solution was then  pipetted into a separate container
          and the sulfate ion (SOO concentration was determined by ion
          chromatography.
                                     24

-------
     4.   The back-half impinger solutions (H?0)  from Runs  3  and  7  were  ana-
          lyzed by ether/chloroform extraction to determine condensible  organ-
          ic content for evaluation of the potential  contribution of  condens-
          ible organics to the difference in reported weights.
     As shown in Table 11, the Method 5B sample fractions exhibited weight
losses relative to their ambient weights ranging  from 30 to 83  percent  in the
probe fraction and 6 to 20 percent in the filter  fraction.  Also  shown  in this
table is a comparison between the ambient Method  5B sample  weights, and  con-
centrations and the within-run Method 5 results reported in Table 8.   For Runs
1, 7, 8, and 9, the Method 5 concentrations were  within 10  percent  of the
within-run Method 5B data versus approximately a  30 percent difference  after
the Method 5B fractions were heated.  For Runs 2, 3,  4, and 5,  ambient  Method
5B concentrations were actually greater than the  within-run Method  5  results.
The ambient Method 5B weights were obtained immediately prior to  heating.
Only one weight was obtained for each fraction after a minimal  desiccation
period.  Some uncombined water may have been retained on these  sample frac-
tions, which could cause a positive bias in ambient weights.   Thus, exact
comparisons between the ambient Method 5B and Method 5 data are probably
distorted to some degree.  The data do show, however, that  condensible  matter
is retained in the front half of the sampling train regardless  of sample
temperature and that the positive effects of the  condensible  matter on  partic-
ulate measurements are reduced by heating the sample fractions  to 160°C.
     As an additional check, sample fractions from Run M5-9 were  heated at
160°C for 6 hours and then cooled and weighed to a constant weight.  The probe
and filter fractions showed relative weight losses of 49 and  20 percent,
respectively.  The Method 5 concentration after heating was 70.0  mg/dNm3
versus 69.5 mg/dNm3 for the within-run Method 5B sample.
     Table 12 presents the results of the residual sulfate  analyses performed
on the rinse and filter fractions for each run.  As shown,  both methods con-
tained a significant amount of water-soluble sulfate (S0.~),  particularly  the
filter fractions.
     Because of limited distillate oil test data  and the fact that  the  par-
ticulate concentrations measured were significantly less than the residual  oil
test data, our discussion on the differences in weights between methods is
limited to the residual oil test data.

                                      25

-------
                               TABLE 11.   METHOD 5B  RELATIVE PERCENT WEIGHT LOSS  AT  160°C
ro
en
Test No.
M5B-1
M5-1
M5B-2
M5-2
M5B-3
M5-3
M5B-4
M5-4
M5B-5
M5-5
MSB-6
M5-6
M5B-7
M5-7
M5B-8
M5-8
MSB -9
M5-9
Ambient temperature3
Sample
weight, mq
Probe
20.4
30.7
23.4
18.2
20.6
16.9
17.4
13.8
22.1
15.8
19.5
57.4
58.0
28.9
34.8
28.7
32.0
Filter
202.8
228.2
123.5
127.7
118.3
100.9
14.4
9.4
4.3
4.1
123.4
261.1
282.7
119.1
111.3
119.3
118.0
Total sample
weight, mg
223.2
258.9
146.9
145.9
138.9
117.8
31.8
23.2
26.4
19.9
142.9
318.5
340.7
148.0
146.1
148.0
150.0
Concen-
tration,
mg/dNm3
70.0
77.5
85.9
82.0
82.7
73.6
16.6
12.2
16.0
12.1
88.8
186.3
199.2
103.5
109.8
90.8
94.9
160°Cb
Sample
weight, mg
Probe
9.1
4.0
6.0
6.3
6.0
6.8
40.3
13.6
12.1
16.2
Filter
169.6
103.2
97.8
13.6
4.0
98.4
211.9
100.4
101.1
94.4
Total sample
weight, mg
178.7
107.2
103.8
19.9
10.0
105.2
252.2
114.0
113.2
110.6
Concen-
tration,
mg/dNm3
56.0
62.7
61.8
10.4
6.1
65.3
147.5
79.7
69.5
70.0
Relative percent
weight loss, %
Probe
55
83
71
64
73
65
30
53
58
49
Filter
16
16
17
6
7
20
19
16
15
20
                aThe Method 5B sample fractions were desiccated  less than 2 hours prior  to heating at 160°C.   One weight
                 was obtained for each fraction.  The Method 5 sample fractions were weighed following procedures
                 described in the Federal  Register (40 CFR 60, Appendix A, July 1984).

                 The reported Method 5B results represented the  sample weights obtained  after heating each fraction  to
                 160°C for 6 hours with a  2-hour desiccation and cooling period per Method 5B.

                cupinht in« - Ambient weight (mq) - 160°C weight (mg)   1QO
                 weignt loss                Ambient (mg)              * luu>

-------
                TABLE 12.   RESIDUAL SULFATE (SO/)  ANALYSIS  OF
                        METHODS 5 AND 5B SAMPLE FRACTIONS
Sample
ID
M5-1
M5B-1
M5-2
M5B-2
M5-3
M5B-3
M5-4
M5B-4
M5-5
M5B-5
M5-6
M5B-6
M5-7
M5B-7
M5-8
M5B-8
M5-9
M5B-9
Sampling
type
Method 5
Method 5B
Method 5
Method 5B
Method 5
Method 5B
Method 5
Method 5B
Method 5
Method 5B
Method 5
Method 5B
Method 5
Method 5B
Method 5
Method 5B
Method 5
Method 5B
Residual sulfate as S0.~,a
Probe
rinse, mg
12.3
7.2
3.2
2.6
2.5
2.0
2.6
2.8
2.5
2.1
2.3
11.4
5.2
5.1
2.8
9.6
6.3
Filter,
mg
105.8
100.9
65.0
55.1
47.8
51.6
3.5
2.3
4.1
2.2
54.7
92.9
67.5
50.8
48.0
57.5
47.3
Total ,
mg
118.1
108.1
68.2
57.7
50.3
53.6
6.1
5.1
6.6
4.3
57.0
104.3
72.7
55.9
50.8
67.1
53.6
Particulate concen-
tration, mg/dNm3
Uncorrected
77.5
56.1
82.1
62.8
75.8
62.0
12.2
10.4
12.1
6.1
65.5
198.9
148.0
109.7
79.9
94.9
69.5
Corrected
42.2
22.1
43.7
28.9
42.2
29.9
9.0
7.7
8,1
3.5
29.9
138.2
105.0
67.8
44.2
52.5
36.6
Analysis by ion chromatography.  Represents total  water-soluble sulfate as SO/

The uncorrected particulate concentrations are as  reported in Table 8.   The
corrected concentrations were calculated by subtracting the quantity of
residual sulfate from the original  sample weights  (Table 8) and dividing by
the appropriate sample volume.
                                    27

-------
     For the residual oil tests, the quantity of water-soluble sulfate found
on the Method 5 samples ranged from 50.3 to 118.1 mg, which represents approx-
imately 31 to 47 percent of the total  particulate collected (Table 8).  The
Method 5B samples exhibited similar characteristics with water-soluble sulfate
quantities ranging from 50.8 to 108.1  mg or approximately 29 to 60 percent of
the total particulate after heating to 160°C.
     Two distinct differences between  the previous studies cited "  and this
project are as follows:
     1.   In the previous studies, all Method 5 samples were subjected to a
          thermogravimetric analysis prior to extraction and determination of
          residual sulfate.  In this study, the Method 5 samples were ex-
          tracted after desiccation at ambient temperature and pressure.
     2.   In the previous studies, the majority of the Method 5 particulate
          catch at ambient conditions  was in the probe rinse and this fraction
          exhibited a significantly higher relative percent weight loss when
          heated to 160°C.  In contrast, the majority of the Method 5 particu-
          late catch at ambient conditions was on the filter fraction.
     As mentioned previously, if HUSO, is the predominant sulfate species in
the gas stream, the Method 5B particulate weights after heating to 160°C
(Table 8) would be expected to be lower than the Method 5 results (which they
are), and correcting for total water-soluble sulfate should account for a
significant portion of the differences in weights (which it does not, Table
12).   Since the quantity of water-soluble sulfate is comparable for each
method, correcting the particulate concentrations from Table 8 for residual
sulfate does not significantly reduce  the difference in results between the
two methods.
     Due to the fact that the Method 5 samples were not heated prior to ex-
traction, the remaining difference in  weights between the two methods (Table
12) may be attributable to water associated with the sulfate in the Method 5
sample.
     The comparability of the Method 5 and 5B SO.  results actually suggests
that hLS04 is not the predominate sulfate species.  If this were the case, the
Method 5B sulfate results should be.considerably lower than the Method 5
results since HUSO* would be significantly reduced by the thermogravimetric
analytical procedure of Method 5B.  The difference in SO." results for Runs 1,
2, 7, 8, and 9 indicate that some HUSO, is present on the sample fraction but

                                      28

-------
the predominate sulfate species are probably metal  sulfates.   This could ex-
plain the difference in participate distribution between the  probe and filter
fractions observed in this study compared to the previous studies cited where
the majority of the particulate and HpSO, was collected in the probe rinse
fraction.  In this study, the particulate catch in  the probe  was small com-
pared to that collected on the filter.
     This test data suggest that the difference in  weights between the two
methods (Table 8) is attributable to H^SO. and its  associated water and/or
water associated with metal sulfates in the particulate.
     The comparability and quantity of the organic  data for Runs 3 and 7
(Table 13) suggest no significant bias resulted from condensible organics.
Considering the sample collection temperature, one  would expect to find a
greater organic content and difference between the  two sample types than
demonstrated here if condensible organics contributed significantly to the
difference in measured concentrations.
       TABLE 13.  ANALYTICAL RESULTS FROM ETHER/CHLOROFORM EXTRACTION
                           OF BACK-HALF SOLUTIONS
Test No.
M5-3
M5B-3
M5-7
M5B-7
Back-half condensible
analysis, mg
Organic
5.9
3.1
4.0
2.6
Inorganic
87.2
73.8
87.5
47.3
Concentration, mg/dNm3
Organic
3.8
1.85
2.3
1.5
Inorganic
56.1
44.1
51.1
27.8
                                      29

-------
                                   SECTION 3
                               QUALITY ASSURANCE

     The objective of testing is to produce representative emission results;
therefore, quality assurance is one of the main facets of stack sampling.
Quality assurance guidelines provide the detailed procedures and actions
necessary for defining and producing acceptable data.   Four such documents
were used in this test program to ensure the collection of acceptable data and
to provide a definition of unacceptable data.  The following documents com-
prised the source-specific test plan prepared by PEI and reviewed by the
Emission Measurement Branch of the EPA; the EPA Quality Assurance Handbook
Volume III, EPA-600/4-77-027; the PEI Environmental  Emission Test Quality
Plan; and the PEI Environmental Laboratory Quality Assurance Plan.  The last
two, which are PEI's general guideline manuals, define the company's standard
operating procedures and are followed by the emission testing and laboratory
groups.
     Relative to this specific test program, the following steps were taken to
ensure that the testing and analytical procedures produced quality data.
     0    Calibration of all field sampling equipment.
     0    Checks of train configuration and calculations.
     0    Onsite quality assurance checks, such as sampling train, pitot tube,
          and Orsat line leak checks and quality assurance checks of all test
          equipment prior to use.
     0    Use of designated analytical equipment and sampling reagents.
     0    Internal and external audits to ensure accuracy in sampling and
          analysis.
     Quality Assurance activities for each specific phase of this project  are
summarized in the following subsections:
                                    30

-------
3.1  CONTINUOUS EMISSION MONITORS
     Each CEM system was set up and operated according to specifications
outlined in the monitor operating manuals.   Performance specifications  (zero
drift, span drift, and response time) outlined in 40 CFR 60,  Appendix B,
Performance Specifications 2 and 3, were followed throughout  this  test  pro-
gram.  Prior to actual stack gas monitoring, a pollutant profile was estab-
lished by traversing the stack cross section and comparing individual sample
point values for NO  and 0/> against a reference point (stack  centroid); this
                   A      £.
permitted a determination of possible gas stratification in the stack.  A
difference of less than 10 percent between  individual sampling points and the
reference data point indicated no significant stratification  problem existed
at the sampling locations.  Tables 14 and 15 show stratification results  for
Boilers 4 and 5, respectively.
     At the beginning of each test day, each monitoring system was leak-
checked and system checks for zero drift, span drift, and response time were
conducted.   The performance specification  tests followed were established for
"continuous on-line" analyzers in operation for long periods  of time.   The
tests applied to monitors in this test series were used as general checks to
ensure reasonable response times and minimal drifts from day-to-day testing.
     Tables 16 through 24 summarize the results of the checks for zero  drift,
span drift, and response time.  The data in Tables 16 through 19 represent
summary data for 24-hour zero and span drift checks.  All drift checks  were
well within the expected operating ranges of the monitors and showed  consist-
ent analyzer response from day-to-day operation.  Response time checks  are
shown in Tables 20 through 24.  All monitors had response times of less than
three minutes for both high-level calibration gas and stack effluent  readings.
Both response times and drift checks show consistent monitor  operation
throughout this test program.
     A three-point calibration was performed on each monitoring system  to
cover the low, mid, and high values of the  specific pollutant concentration
measured.  This system check was conducted  at the beginning and end of  most
test blocks.  Single-point calibration checks were performed  between  test
blocks when sufficient time was not available for a three-point calibration
check.  Calibration gases were transported  through the sample-conditioning

                                     31

-------
           TABLE 14.   MONITOR  STRATIFICATION TEST—BOILER 4 (8/14/84)

Traverse
Point No.
West Port
1
2
3
4
North Port
1
2
3
4

NO
concentra-
tion, ppm
_
133
132
138
142
_
140
142
141
143
Reference
NO con-
centra-
tion, ppm
134
132
132
136
142
140
142
141
141
146

NO Devia-
tion, %
_
+0.75
0
+1.5
0
_
-1.4
+0.71
0
-2.1

02 con-
centra-
tion, %
_
3.2
3.2
3.1
3.3
_
3.2
3.1
3.2
3.2
Reference
02 con-
centra-
tion, %
3.2
3.2
3.2
3.3
3.2
3.1
3.1
3.1
3.1
3.0

02 Devia-
tion,0 %
_ i
0
0
-6.1
+3.1
^
+3.2
0
+3.2
+6.7
 Reference point is  the  sampling point located in the center of the  sample matrix.
•Went deviation . Tr"erie^g.^gnce- PP" x 100.
           TABLE 15.   MONITOR STRATIFICATION TEST—BOILER 5 (8/16/84)


Traverse
Point No.
West Port
1
2
3
4
North Port
1
2
3
4

NO
concentra-
tion, ppm
_
352
347
342
329
.
342
342
341
327
Reference
NO con-
centra-
tion, ppm
352
352
352
345
331
337
347
347
339
327


NO Devia-
tion, %
_
0
-1.4
-0.9
-0.6
.
+1.5
-1.4
+0.6
0

02 con-
centra-
tion, %
_
1.5
1.3
1.3
1.6
^
3.6
3.5
3.5
3.5
Reference
02 con-
centra-
tion, %
1.3
1.3
1.3
1.3
1.2
3.5
3.6
3.5
3.5
3.5


02 Devia-
tion,0 %
_
+15.4
0
0
+25
^
0
0
0
0
  eference point  is  the  sampling point located in the center of the  sample matrix,
ercent deviation .
                                               ""
                                                   x 100.
                                    32

-------
                         TABLE 16.  TEST RESULTS FOR NO  MONITOR 24-HOUR ZERO AND CALIBRATION  DRIFT
                                              (ppm NO  except as indicated)
                                                     X
Test No.
1
2
3
4
5


Date (1984)
Start
8/11
8/14
8/16
8/17
8/18
End
8/12
8/15
8/17
8/18
8/19
Test time
Start
1215
0800
1500
0830
0800
End
0930
0800
1500
0800
0900
Zero reading
Start End
(A)
-0.5
2.4
2.4
0.7
0.5
I(B)
0
0
0.7
0.5
0
Arithmetic mean (AM)
95% confidence interval (CIg5)
24-hour drift,3 %
Zero
drift
(C=B-A)
0.5
-2.4
-1.7
-0.2
-0.5
0.9
1.35
0.23
Span reading
Start End
(D)
222
227
426
441
470
IE)
225
223
444
470
427


Span
drift
(F=E-D)
3
-4
18
29
-43


Cali-
bration
drift
(F-C)
2.5
1.6
19.7
29.2
-42.5
19.1
34.2
5.3
co
co
24-hour drift =
                              CI
                           1QQO
                                    x 100.
      Zero drift
                                                           Calibration  drift
              CIQ, =  2'776   x|  5 (9.19) - (18.5) = 1.35
                yb    5 xTT
                                                           95
                                                              =   2-776     J  5  (3056)  -  (110.3)  =  34.2
                                                                    ~~

-------
                        TABLE 17.   TEST RESULTS FOR CO MONITOR 24-HOUR ZERO AND CALIBRATION DRIFT

                                             (ppm CO except as indicated)



Test No.
1
2
3
4
5
6
7





Date
Start
8/11
8/12
8/13
8/14
8/15
8/17
8/18
1984)
End
8/12
8/13
8/14
8/15
8/16
8/18
8/19


Test time
Start
1215
0930
0820
0800
0800
0830
0800
End
0930
0820
0800
0800
1100
0800
1940

Zero reading
Start End
(A)
0
-0.2
0
0.5
-0.2
0
0.5
(B)
-0.2
0
0.5
-0.2
0
0.5
0
Arithmetic mean (AM)
95% confidence interval (CIg5)
24-hour drift,9 %

Zero
drift
(C=B-A)
-0.2
0.2
0.5
-0.7
0.2
0.5
-0.5
0.4
0.44
0.40

Span reading
Start End
(D)
202
201
204
203
201
203
206
(E)
201
204
203
201
203
206
204




Span
drift
(F=E-D)
-1
3
-1
-2
2
3
-2



Cali-
bration
drift
(F-C)
-0.8
2.8
-1.5
-1.3
1.8
2.5
-1.5
1.74
1.77
1.14
co
-P.
a% 24-hour drift
Zero drift
CI95 =
|AM| + CI95
x 100

ZTtfc 1
• \l 7 ( 1 36) (0 04) ~ 0 44
7 xT6~
Calibration drift

rT _ 2.365 1 7
cm , 	 N '
yo 7 c
/ \^ b

(24.16) - (4) = 1.77

-------
                        TABLE 18.  TEST RESULTS FOR 02 MONITOR 24-HOUR ZERO AND CALIBRATION  DRIFT
                                              (% 02 except as indicated)
Test No.
1
2
3
4
5
6
7


Date
Start
8/11
8/12
8/13
8/14
8/16
8/17
8/18
(1984)
End
8/12
8/13
8/14
8/15
8/17
8/18
8/19
Test time
Start
1215
0930
0820
0800
1500
0830
0800
End
0930
0820
0800
0800
1500
0800
0900
Low-range
reading
Start End
(A)
0.99
0.96
0.94
0.96
1.02
1.00
0.97
(B)
0.96
0.94
0.96
0.97
0.99
0.97
0.97
Arithmetic mean (AM)
95% confidence interval (CIg5)
24-hour drift, b %
Low-
range
drift
(C=B-A)
-0.03
-0.02
0.02
0.01
-0.03
-0.03
0
0.02
0.02
0.04
Span reading
Start End
(D)
8.08
8.05
8.01
8.03
7.93
7.93
7.70
(E)
8.05
8.01
8.03
8.04
7.93
7.70
7.79


Span
drift
(F=E-D)
-0.03
-0.04
0.02
0.01
0
-0.23
0.09


Cali-
bration
drift
(F-C)
0
-0.2
0
0
0.3
-0.2
0.9
0.05
0.08
0.13
CO
in
   aThe low-range calibration gas (1.003 % 02) data were substituted for the zero drift check.   The data test
    monitor is set up on a calibration gas basis (no actual zero); zero readings are dependent  on the low-range
    calibration gas used.

    \ 24-hour drift as % 02 =  |AM| + CIg5.
     Zero drift
  Calibration drift
             CIQ, =  2'365    J 7 (0.0036) - (0.006) = 0.02
               yb    7\|  6
CI
        2.365
  95
\ 7  (0.05)  -  (0.01)  = 0.08

-------
                      TABLE 19.  TEST RESULTS FOR C02 MONITOR 24-HOUR ZERO AND CALIBRATION DRIFT
                                             (% C02 except as indicated)
Test No.
1
2
3
4
5
6
7



Date
Start
8/11
8/12
8/13
8/14
8/16
8/17
8/18
[1984)
End
8/12
8/13
8/14
8/15
8/17
8/18
8/19
Test time
Start
1215
0930
0820
0800
1500
1500
0800
End
0930
0830
0800
0800
1500
1800
0900
Zero reading
Start End
(A)
0
0.04
0
0
0
0.10
0
(B)
0.04
0
0
0.02
0.10
0.10
0
Arithmetic mean (AM)
95% confidence interval (CIg5)
24-hour drift,3 %
Zero
range
drift
(C=B-A)
0.04
-0.04
0
0.02
0.10
0
0
0.03
0.04
0.07
Span reading
Start End
(D)
15.4
15.8
16.3
15.9
14.7
15.1
15.6
(E)
15.8
16.3
15.9
15.5
15.1
15.6
15.7



Span
drift
(F=E-D)
0.40
0.50
-0.40
-0.40
0.40
0.50
0.10



Cali-
bration
drift
(F-C)
0.36
0.54
-0.40
-0.42
0.30
0.50
0.10
0.37
0.36
0.73
GO
CTl
a% 24-hour drift
Zero drift

qc -
as % C02 = |AM| + CI95.
2Oi;c i—
.JD3 y /Q Q136) (Q 01
7 r~r N 	 '
Cal ibration drift

nn\ n 04 TT - ^•JDO 1 7
yb 7 J~l~

(1.11) - (0.96) = 0.36

-------
          TABLE  20.  MONITOR  RESPONSE TIME
                       (seconds)
Test No.
NO
°2
CO
co2
Upscale9 at
1110 (8/17/84)
2:00
2:00
2:45
2:30
Downscale
2:10
2:20
2:50
2:15
 Response  time  needed  to  record  stable  stack effluent
 reading.

Response  time  needed  to  record  stable  high-level calv
 bration gas  reading.
         TABLE  21.   NO  MONITOR  RESPONSE TIME
                       (seconds)
Test No.
1
2
3
Date
(1984)
8/12
8/18
8/19
Downscale9
1:58.9
2:05
2:00
 Response  time  needed  to  record  stable  high-level  calv
 bration gas  reading.
                           37

-------
         TABLE 22.  02 MONITOR RESPONSE TIME
                      (seconds)
Test No.
1
2
3
Date
(1984)
8/13
8/13
8/14
Downscale9
2:15.0
2:30.0
2:15
Response time needed to record stable high-level  cali-
bration gas reading.
         TABLE 23.  CO MONITOR RESPONSE TIME
                      (seconds)
Test No.
1
2
3
Date
(1984)
8/12
8/13
8/14
Downscale
3:08.6
2:48.1
2:30
Response time needed to record stable high-level calv
bration gas reading.
        TABLE 24.  C02 MONITOR RESPONSE TIME
                      (seconds)
Test No.
1
2
3
Date
(1984)
8/12
8/13
8/14
Downscale3
2:17.5
1:54.3
2:00
Response time needed to record stable high-level cali-
bration gas reading.
                           38

-------
system and sample line as a system check and an indicator of possible sample
dilution or contamination.  All calibration gases were Master Gas-Certified,
which means the gas values were within ±2 percent of indicated values.   Along
with calibration gases, zero nitrogen was used to zero all  monitors  and to
purge sample lines to guarantee a clean sampling system.
     Data generated by the CEM calibrations (three-point  and single-point)
were used to define calibration curves for each monitoring system.   Each
calibration response had a chart division reading and a corresponding cali-
bration gas concentration (ppm, %).  A linear regression  analysis of these
data was conducted to establish the relationship between  response and concen-
tration, or, the degree of correlation (linearity).   Figures 1 through 4
present example calibration curves.  The linear regression equations estab-
lished for each monitor on a daily basis were then used to define pollutant
concentrations for each specific test block.  The final reduction of data was
accomplished by taking an average chart reading for every 5-minute period and
determining the pollutant concentration by the linear regression equation.
Tables 25 through 28 summarize the CEM linear regression  data for each test
block conducted.
     The EPA supplied NO  audit gases to check monitor response and  accuracy.
                        A
These audit gases were analyzed daily throughout the test program.   Table 29
summarizes the results of the NO  CEM system audits.  As  shown, the  NO  CEM
                                A                                     A
response compared favorably with the audit cylinder values.
     As a final check of the NO  CEM system, several stack samples were col-
                               A
lected and analyzed according to procedures described in  EPA Reference Method
7.*  Table 30 summarizes the comparative data.  Results for the majority of
the Method 7 samples collected were within ±20 percent of the NO  CEM values
                                                                A
recorded during the sample collection period.
     Table 31 summarizes a data comparison between 02 and CO,, results from
those test blocks for which both CEM and EPA Reference Method 3 data are
available.  The CEM data represent average values for each designated test
block.  The Reference Method 3 data were obtained by collecting a gas sample
in a Tedlar bag during the particulate tests.  The gas sample collection probe
was attached to a particulate train sample probe during each test so that an
 40 CFR 60, Appendix A, Reference Method 7, July 1984.
                                      39

-------
                   NOV CEM CONCENTRATION, 8/15/84,
                           TEST BLOCK 1-G
                                XPPM  0.425
                   CORRELATION COEFFICIENT =  0.9994
           120      160      200
          NO CONCENTRATION, PPM
Figure 1.  Example NOX calibration curve.
                40

-------
  100
   90
   80
   70
   60
    50
eC
U
    40
    30
    20
    10
09 CEM CONCENTRATION,  8/15/84,
 *     TEST BLOCK 1-G


         u Y-4.41
                                                      10.18
                                                         CD-4.41
                                                          10.18
                                         CORRELATION  COEFFICIENT  =  0.9998
                                                    10
                  12       14
                   Figure  2.  Example 02 calibration curve.

                                     41

-------
o
   100
    90
    80
    70
    60
     50
    40   •
    30   -
    20   -
     10   -
                40
                   CO  CEM CALIBRATION, 8/15/84,
                        TEST BLOCK 1-G

                         v.Y-5.51
                         *  0,404
                                                   CO
                               _CD-5.51
                            PPM  0.404
                                          CORRELATION COEFFICIENT = 0.9998
80       1ZO     160       200

               C02,PPM
240
280
                   Figure  3.  Example CO calibration curve.
                                      42

-------
   100

   Y


    90




    80



    70




    60
o
•—H
t—I
>•
S  50
i—
or
3C

    40



    30



    20




    10
                         C02 CEM CALIBRATION, 8/15/84,
                               TEST  BLOCK  1-G
                                  X=
Y-6.08
4.856
                                  rn y-CD-6.08
                                   V  4.856

                     CORRELATION  COEFFICIENT = 0.9988
     6        9        12       15       18
                    co2,%

Figure 4.   Example C02 calibration curve.
                43
              21

-------
                     TABLE 25.
NO  LINEAR REGRESSION DATA
  /\
Test Block
l-Ga
2-G
lb
2
3
4
5
6
7
8
9
3-6
4-6
5-6
6-6
7-6
8-6, 9-6
and 10-6
No. of cali-
bration points
7
7
8
8
8
8
8
8
6
8
8
8
6
6
6
4
6
Y-intercept
6.80
7.68
5.90
5.84
5.55
5.41
5.40
5.13
6.07
5.84
5.60
5.01
5.02
5.01
4.98
5.40
5.24
Slope
0.4252
0.4125
0.0994
0.0975
0.0970
0.1027
0.1050
0.1070
0.0950
0.1013
0.1038
0.1009
0.1020
0.1032
0.1047
0.1057
0.1044
Correlation
coefficient
0.9994
0.9994
0.9996
1.0000
0.9998
0.9999
0.9996
0.9998
0.9995
0.9954
0.9978
0.9998
1.0000
0.9999
0.9999
0.9995
0.9996
JTest Blocks  1-6 and 2-G on 0-250 scale.


5A11  other test blocks on 0-1000 scale.
                                      44

-------
TABLE 26.  0, LINEAR REGRESSION DATA
Test Block
1-6
2-6
1
2
3
4
5
6
7
8
9
3-6
4-6
5-6
6-6
7-6
8-6, 9-6
and 10-6
No. of cali-
bration points
5
5
6
6
6
6
6
6
4
6
6
6
4
4
5
4
5
Y-intercept
4.41
4.82
7.39
7.24
7.22
4.78
4.68
4.92
5.04
4.85
5.37
4.58
4.53
4.44
3.64
1.87
3.33
Slope
10.18
9.97
9.90
9.67
9.70
10.03
9.88
9.14
8.78
8.88
9.00
9.88
9.90
9.62
9.82
10.20
9.91
Correlation
coefficient
0.9998
0.9998
0.9994
0.9999
0.9998
1.0000
0.9994
0.9944
0.9991
0.9994
0.9999
0.9999
0.9999
0.9999
0.9997
1.0000
0.9995
                 45

-------
                    TABLE 27.  CO LINEAR REGRESSION DATA
Test Block
1-G
2-6
1
2
3
4
5
5a
6
7
8
9
3-6
4-6
5-6
6-6
7-6
8-6, 9-6
and 10-G
No. of cali-
bration points
7
7
8
8
8
8
8
4
8
6
8
8
8
6
6
6
4
5
Y-intercept
5.51
5.95
5.46
5.60
5.49
5.45
5.65
6.34
5.56
6.84
6.76
6.36
5.32
5.08
5.09
5.17
4.95
4.97
Slope
0.4039
0.4028 .
0.4061
0.4063
0.4068
0.4064
0.4066
0.1857
0.4072
0.4068
0.4064
0.4084
0.4052
0.4065
0.4058
0.4075
0.4089
0.4082
Correlation
coefficient
0.9998
0.9999
0.9999
0.9999
0.9999
0.9999
1.0000
0.9963
0.9999
0.9998
0.9999
0.9999
0.9999
0.9999
0.9999
0.9999
1.0000
1.0000
0-500 ppm scale; all other tests 0-250 ppm scale.
                                     46

-------
TABLE 28.
C02 LINEAR REGRESSION DATA
Test Block
1-6
2-6
1
2
3
4
5
6
7
8
9
3-6
4-6
5-6
6-6
7-6
8-6, 9-6
and 10-6
No. of cali-
bration points
7
7
8
8
8
8
8
8
6
8
8
8
6
6
6
4
6
Y-intercept
6.11
6.10
5.55
5.85
6.15
6.54
6.94
6.65
6.14
6.99
6.79
6.30
5.98
6.28
5.89
5.45
5.52
Slope
4.85
4.79
4.70
4.73
4.88
4.88
4.82
4.76
4.88
4.83
4.94
4.93
4.77
4.76
4.97
5.14
4.99
Correlation
coefficient
0.9988
0.9994
0.9998
0.9996
0.9985
0.9993
0.9988
0.9982
0.9993
0.9991
0.9995
0.9987
0.9995
0.9992
0.9990
1.0000
0.9990
                 47

-------
                    TABLE 29.  SUMMARY  OF  NO   OEM AUDIT  RESULTS
Date
(1984)
8/11
8/11
8/14
8/15
8/16
8/16
8/17
8/17
8/18
8/18
8/18
8/18
8/18
8/19
8/19
EPA audit
cylinder ID
LL-4348
AAL-6222
LL-4348
LL-4348
AAL-6222
AAL-6222
AAL-6222
LL-4348
AAL-6222
AAL-6222 '
AAL-6222
AAL-6222
AAL-6222
AAL-6222
LL-4348
EPA audit .
value NO, ppm
109
225.2
109
109
225.2
225.2
225.2
109
225.2
225.2
225.2
225.2
225.2
225.2
109
CEM NOC
value, ppm
113
225
110
113
223
226
222
109
215
225
220
238
223
223
110
Percent
difference
+3.5
-0.09
+0.90
+3.5
-0.99
+0.35
-1.4
0
-4.7
-0.09
-2.4
+5.4
-0.99
-0.99
+0.90
aGas cylinders provided by U.S.  EPA.

 Audit values of nitric oxide (NO) with the balance of gas being nitrogen.

cValues determined from PEDCo NOX CEM system.   All  audit gases were introduced
 at the sample probe.
                                      48

-------
           TABLE 30.  COMPARISON  OF  REFERENCE  METHOD  7 AND NO
                           CEM TEST RESULTS                 >
Method 7
Test No.
G-1A
G-1B
G-1C
G-1D
1A
IB
1C
ID
7A
7B
7C
8B
8C
G-6A
G-6B
G-6C
G-6D
Date
(1984)
8/15
8/15
8/15
8/15
8/16
8/16
8/16
8/16
8/18
8/18
8/18
8/18
8/18
8/19
8/19
8/19
8/19
Time
(24-h)
1642
1650
1658
1704
1350
1400
1914
1920
1756
1805
1815
2125
2130
1826
1831
1835
.1837
Testing
location
No. 4
No. 4
No. 4
No. 4
No. 5
No. 5
No. 5
No. 5
No. 5
No. 5
No. 5
No. 5
No. 5
No. 4
No. 4
No. 4
No. 4
Method 7
test
result, ppm
151
161
151
160
379
404
348
368
331
363
359
374
333
138
116
134
122
CEM NO
value, ppm
193
194
196
199
361
361
381
381
430
430
433
394
394
140
120
124
124
Percent
difference
+28
+20
+30
+24
-5
-11
+9
+4
+30
+18
+21
+5
+13
+1.4
+3.4
-7.5
+1.6
Difference =
CEM - Method 7
   Method 7
x 100.
                                    49

-------
          TABLE 31.  COMPARISON OF OXYGEN AND CARBON DIOXIDE
             RESULTS—CEM AND  REFERENCE METHOD 3  (ORSAT)
Test Block
1
2
5
8
Date
(1984)
8/16
8/16
8/17
8/18
CEM value9
02
3.5
1.0
1.2
1.4
C02
13.2
15.0
15.6
15.4
Reference Method 3
02
4.6
2.3
2.0
2.5
C02
12.6
13.7
13.5
13.5
 Represents  average  monitor  value  calculated for the designated
 test  block.

'Represents  data  from  Method 3  analysis  (Orsat) of  integrated bag
 samples  collected during  the designated  particulate tests.
                                50

-------
integrated (traverse) sample was collected.   No attempt was  made  to collect
gas samples at a point near the CEM sample probe.
     Several  Method 3 samples (Test Blocks 3, 4, 6, and 7) were considered
void because  of broken bags or leakage problems within the gas  sampling  sys-
tem.  Accordingly, the CEM data from these test blocks were  used  in all  par-
ticulate calculations.  As presented in Table 31,  the EPA Reference Method 3
data consistently exhibited higher CL and lower CCL values than the average
CEM results.   Though no specific explanation was found for the  discrepancy,
the Method 3  samples might have been diluted during sample collection or
immediately prior to analysis.  Considering  the low levels of CL  encountered,
such a dilution would tend to distort the comparison between the  two measure-
ment systems.  In contrast, the difference between the Method 3 and CEM  C02
values ranged between 5 and 15 percent, which is considered  acceptable.

3.2  MANUAL TESTS--PARTICULATE AND NO
                                     X
     Table 32 lists the sampling equipment used to perform the  particulate
tests and the calibration guidelines and limits.  In addition to  the pre- and
post-test calibrations, a field audit was performed on the metering systems
and thermocouple digital indicators used for sampling.  Critical  orifices
constructed by PEI were used in the dry gas  meter audits. Figures 5 through  7
present results of the onsite audits.  These data were used  to  assess the
operational status of the sampling equipment relative to EPA guidelines.
     Figure 8 is an example of an unacceptable meter box audit.  The audit
value for the dry gas meter Y-factor was greater than ±5 percent, which  was
considered unacceptable by the PEI Project Manager; therefore,  the meter box
was not used  for this test program.
     The sample data and isokinetic sample rates were calculated  on site by
PEI personnel.  The data were rechecked and  validated at the end  of the  test
program by computer programming.
     Figure 9 presents an example calculation form PEI used  during the test
series.  Computerized calculations are presented in Appendix A  of this report.
     As a check of the analytical methodology used, blank filter and reagent
(acetone) samples were analyzed in a similar fashion as the  actual field
sample analyses.  Table 33 presents the gravimetric blank analysis blank.
                                     51

-------
                                   TABLE  32.  FIELD EQUIPMENT CALIBRATION
01
ro
Equipment
Meter box

Pilot tube
Digital indi-
cator
Thermocouple
and stack
thermometer
Orsat analyzer
Inpinger
thermometer
Trip balance
Barometer
Dry gas
thermometer

Probe nozzle
ID
Mo.
FB-4
FB-8
188
515
126
221
201
205
0,
c62
CO
385
291
M-l
229
FB-4
FB-8
8-107
8-110
6-109
6-110
Calibrated
against
Wet test meter

Standard pitot
tube
Millivolt signals
ASTM-2F or 3F
Standard gas
ASTM-2F or 3F
Type S weights
NBS traceable
barometer
ASTM-2F or 3F

Caliper
Allowable
error
Y ±0.02 Y
AH @ iO.15
(Y ±0.05 Y post-test)

Cp *0.01
0.5*
1.5%
(±2* saturated)
±0.5%
±2°F
±0.5 g
±0.10 in.Hg
(0.20 post-test)
±5°F

Dn ±0.004 in.
Actual
error
0.67
0.05
-0.68
0.02
0.05
-2.4
0.0
<0.5%
<0.5l
<0.50%

-------
                                  ON-SITE AUDIT DATA SHEET
          Audit Name:
Date:  g///AV   Auditor:
    r*
TW>.
 I*
 +*•
Equipment
'-V Meter box
.^ inlet thermo.
-/ Meter box
-i. outlet thermo.
"^ Impinger
•6 thermometer
„ Stack
thermometer
or
Thermocouple
Orsat
analyzer
Trip
balance
Barometer
Reference
ASTM-3F at
ambient temp.
ASTM-3F at
ambient temp.
ASTM-3F at
ambient temp.
ASTM-3F at
ambient temp.
ASTM-3F at
stack temp.
% 02 in
ambient air
IOLM std.
weight
Corrected*
NWS value
Reference
Value
•^yf
W
77-
7>V^
I*'?
20.8%
ii
^
Value
Determined
7/V
73 V
e*°
Wf
irf
>fl.«


Deviation
Z>
-7£
^3*^
-*•/*
-/'?
0,1


Max. Allowable
Deviation
5°F
5°F
2°F
7°F
See table
0.7%
0.5 grams
0.20 in. Hg
Reference temp. °F
Max. deviation °F
32-140
7
141-273
9
274-406
11
407-540
13
541-673
15
674-760
17
           *  Correction factor:

              NWS value (in.  Hg) - [Altitude  (ft)/1000(ft/in.  Hg)] + 0.74 in.  Hg**

          **  0.74 1n. Hg is the nominal correction factor for the reference barometer
             against which the field barometer was calibrated.

          If  It  Is not feasible to perform the audit on any piece of equipment, record
          "N/A"  In the space provided for the  data.
                                                     Iff
                                      It'F
                                      I
                                                      Iff
                     -3'f
                    Figure  5.   On-site audit  data  sheet.

                                               53

-------
                      FIELD AUDIT REPORT:  DRY GAS METER
                             BY CRITICAL ORIFICE
DATE:
                CLIENT:
BAROMETRIC  PRESSURE (Pbjr):^T.79 in.Hg  METER BOX NO.

ORIFICE NO.       3	  PRETEST Y: ^

ORIFICE K FACTOR:    5^3">7X /*'v     AUDITOR:
                                     AHP
                                                                      1n.H20

Orifice
nanometer
reading
AH,
in.HzO
/.^
Dry gas
meter
reading
w
ft3
1i.ee>
f*-lfa_i
Temperatures
Ambient
Tai/Taf
°F
1M-
73
Average
V
°F
13*
Dry gas meter
Inlet
VTif
°F
?>
°ll
Outlet
Toi/Tof
°F
n
VB
Average
Tm'
"F
g6.7^
Duration
of
run
0
min.
/jTosy
Dry gas
meter
V fts
/***
Vm
mstd'
ft3
/*.*1
Vm
macf
ft3
/^.^

Audit,
Y
/..»/
Y
devia-
tion, %
'•>•<•&
Audit
AH(?,
1n.H20
) /./^

AHP Devia-
tion, 1n.H20
& - . o^ (p4-s
 "std
            17.647(Vj(Pbar + AH/13.6)
460)
                                                ft3
 "act
            1203( 0 )( K ){Pbar)
              (Ta + 460)
                        112
                                                ft3
Audit Y
             "'act
             "'std
      Y  deviation
                                           Audit Y - Pre-test Y
                                                  Audit Y
x 100
Audit AH?  =  (0.0317)(AH)(P.aJ{T  + 460)
                         DO r  m
Audit Y must be in the range,  pre-test Y ±0.05 Y.
Audit AHP  must be in the range pre-test AH@ ±0.15  inches H20.
        Figure  6.  Field  audit of  dry  gas meter  (Meter  Box  FB-4)
                                      54

-------
                     FIELD AUDIT REPORT:  DRY GAS METER
                             BY CRITICAL ORIFICE
DATE:
                                      CLIENT:
BAROMETRIC PRESSURE (P^):/??.?? 1n.Hg  METER BOX NO.
ORIFICE NO.
                                      PRETEST Y:
                                                          AHP  1.11   1n.H20

Orifice
manometer
reading
AH,
1n.H20
' '9 ""V
Dry gas
meter
reading
VV
ft3
700.*
7*t.±
Temperatures
Ambient
Tai/Taf
°F
77
75-
Average
V
°F
^
Dr
Inlet
VTif
°F
ev
&<£,
1 gas meter
Outlet
w
°F
77
7^
Average
Tm-
°F
e,r
Duration
of
run
0
min.
*:«*
Dry gas
meter
V ft3
/,?.7^
mstd-
ft3
/P.VV
macf
ft3
/p.sr
Audit,
Y
/.009
Y
devia-
tion, X
/-/.^^
Audit
AH@,
1n.H20
/.e
-------
                      FIELD AUDIT REPORT:   DRY GAS METER
                             BY CRITICAL  ORIFICE
DATE:
                                       CLIENT:
BAROMETRIC  PRESSURE (Pbar):.*?.S7 in.Hg  METER BOX NO.

ORIFICE  NO.       3	  PRETEST Y: _^

ORIFICE  K FACTOR:   j'377*/l>~'f       AUDITOR:  C
                                                           AHP  A7V  1n.H20
(\
Orifice
manometer
reading
AH,
in.HjO
s
» £"/**
Dry gas
meter
reading
VV
ft3
Jtf. 4o
3*x*>
Temperatures
Ambient
Tai/Taf

°F
16
11-
Average
Ta'
°F
^
7.i
Dry gas meter
Inlet
T11/T1f
°F
73
7/
Outlet
Toi/Tof
°F
73-
73
Average
Tm-
°F
72

Duration
of
run
min.

//"•<*

Dry gas
meter
Vm- ft3
/;.y
Vm
"std1
ft3
IM
Vm
macf
ft3
/3,-)l

Audit.
Y
,.»11
Y
devia-
tion, %
'-'ft
Audit
AHP,
in.H20


AH@ Devia-
tion, in.H20

 "std
 "act
            1203( 0 )( K )(Pbar)
              (Tfl 4 460)
                        \Ti
                                          #*
                                               ft3
                                         fi.Tl
AuditY=
                             Y deviation *
                                           Audit Y
             std
                                                                   lo° g ^'
Audit
            (0.0317)(AH)(Pbar)(Tm + 460)    (V(t,    , AH/l3.6
                                                                        in.H20
Audit Y must be in the range,  pre-test Y ±0.05 Y.
Audit AHP must be in the range pre-test AH@ ±0.15  inches H20.
         Figure 8.   Example of  an unacceptable meter  box  audit.
                                     56

-------
                                                    14
                                    ISOKINETIC CALCULATION
 SITE.
        fi/* $
                              -if
                                                            RUN  1
                                                                 RUN 2
                                                                               RUN 3
RUN 4
 1.
VeliMi of dry fit tM^lrt cormttd to
tundtrd condition.  HeU: »„ mitt kt
Cormtld for iMUgt 1f t*J )l*k«at
ntet fiCMd l().


    • I7.M » »_ « T  tar • TIT
                                            V..  ft3
                                                                    Mf A ' Z.
                                            AH. 1n.H20
                                                        0.7
                                                                      .7
                                            V    . dscf
                                             "$td

2.  Voluat of Mttr »«por tt tUndcrd con-
    d1t1en». ft'.
                                       Vlc. 9
     Vd
          0.04707V
                 ..
                 *e
                                             .ft
                                                                     to,(ej
                                                                           1-74
3.  ItoiiUM content in lUek 9ts.
    •«•
            •ttd
          V     • V
          •»td    *
4.  Dry •oltculir M * » "« *
                                       Ms. Ib/lb-nole
                                                                .
                                                            x. a, . / >«|-
                                                            ^- / • 7 U
(.   Stack *t1oc1tjr at tuck condition*.
    »P».
                                            Pstat1c* 1n
-------
TABLE 33. PARTICULATE FILTER AND REAGENT BLANK ANALYSIS
Sample type and
identification
Method 5 - ambient
Filter No. 0002771
(D5079)
Acetone (D5088)
(Blank volume
= 170 ml)
Method 5B - 160°C
Filter No. 0003396
Acetone D5088
(Blank volume
= 170 ml)
Original tare
weight, mg
347.5
101,882.1
349.5
106814.0
Blank
weight, mg
347.7
101,888.8
349.6
106815.1
Net
weight, mg
+0.2
+6.8 mg
= 0.051 mg/g
+0.1
+1.1
                         58

-------
     Figures 10 through 13 present the results of the NO  (Method 7) audit
                                                        /\
results from both the field and laboratory analyses.  Audit solutions supplied
by EPA were analyzed according to the procedures described in Method 7.  The
results indicate good analytical  technique.
                                     59

-------
         AUDIT REPORT NOV  ANALYSIS
Plant
(?,£,
PN Number
Date sample
Samples ana
Reviewed by

Sample
Number
3SJ3





s received & -/£
lyzed by //• L ,
7/2-,
f
mg N02/dscm
Determined
^~ 3 % ?





'-?*/ Date a

Date o

Source of
Sample
r*-^





nalyzed

f Review

Accepted
Value
W.9





?-/*-??

S/3//9"/

Difference
-9.V





Figure 10.  Audit report NO  analysis results  on samples
                 received 8/15/84.
                       60

-------
                 AUDIT REPORT NOX ANALYSIS
Plant
PN Number
                                                          7
Date samples received   <^-/

Samples  analyzed by    tf> /.

Reviewed by
                                     Date  analyzed
                                     Date  of Review
Sample
Number
LOf OH';!*,
yj?3





mg N02/dscm
Determined
$J3





Source of
Sample
7" fay^^





Accepted
Value
93 7, A





%
Difference
+ /0*3





         Figure 11.  Audit report NO  analysis results on samples
                         received*8/16/84.
                                61

-------
Plant
                  AUDIT REPORT  NOX ANALYSIS
PN Number
Date samples received y-/9-fY. ?-/^/Date analyzed P'/f-P^
Samples analyzed by /^«//
Reviewed by
Sample
Number
* S°SA





'%/.%'. Date of Review %/?//f"/
// f f
mg NC>2/dscm
Determined
/£>/&





Source of
Sample
r.*r~





Accepted
Value
f«*





Difference
.*?





       Figure 12.  Audit report NO  analysis results  on samples
                  received  8/17/84 and 8/18/84.
                                 62

-------
        AUDIT REPORT NOV ANALYSIS
Plant    £;£  6<,'(/6.//l  ~   (JtCPii

Date samples received	

Samples  analyzed by   &

Reviewed by	/T/. a---
                             PN Number
                             Date analyzed
                             Date of Review   % '5,'/
Sample
Number
3 *n -~}
e>(j I t-1-
'R5o^iT
?£??
T^o^t*




mg N02/dscm
Determined
3/f.5"^
^7 U ° «r
/ 7^
-------
                                   SECTION 4
                      SAMPLING LOCATIONS AND TEST METHODS

4.1  SAMPLING LOCATIONS
     All tests were run in the exit stacks of Boilers 4 and 5, as depicted in
Figures 14 and 15.  In the Boiler 4 exit stack, two sampling ports, 90 degrees
off center, were located approximately 3.2 duct diameters downstream and more
than 8 duct diameters upstream from the nearest flow disturbance in the 152-cm
(60-in.) i.d. round stack.  A total of 24 traverse points (12 per port) were
used to measure gas velocity and temperature.  As mentioned in Section 2, no
particulate measurements were made in this stack.
     In the Boiler 5 exit stack, four sampling ports, 90 degrees off center,
were located approximately 2.7 duct diameters downstream and more than 8 duct
diameters upstream from the nearest flow disturbance in the 183-cm (72-in.)
i.d. round stack.  A total of 24 sampling points (12 per port) were used to
conduct the particulate tests.  Two individual sampling trains (a Method 5
with a sample collection temperature of 120°C and a Method 5B with a sample
collection temperature of 160°C) were used to isokinetically traverse the
cross-sectional area of the stack for each of nine test runs.  During Tests  1
through 5, each point was sampled for 5 minutes, for a total test time of 120
minutes.  During the remaining tests, the total test time was reduced to 72
minutes, or 3 minutes per sampling point.  A brief description of the test and
analytical procedures used is presented in the following subsection.

4.2  CONTINUOUS EMISSION MONITORS—SAMPLE EXTRACTION, ANALYSIS, AND DATA
     REDUCTION
     An extractive monitoring system was assembled on site to provide a con-
tinuous emissions data base for NO , 02, CO, and C02-  Figure 16 presents the
CEM system layout.
                                     64

-------
                     TO ATMOSPHERE
      SAMPLE PORTS
      10-cm  (4-in.)
          I.D.
ROOF  LINE
                        152 cm
                     (60-in.) I.D.
                       02 MONITOR
                                                 CROSS-SECTION
                                                 MONITOR LOCATION
    3.7 m
(12  ft.  2 in.)
                                                     4.9 m
                                                    (16 ft.)
    1.2 m
(»3  ft. 10 in.)
                        FLOW FROM
                         BOILER
        Figure  14. Boiler  4 exit stack (no scale).
                           65

-------
/* r
t C
PORT
LOCATION
10-cm (4-in.)
l.D.
ROOF LINE

_ 183 cm 	 „
(72 in.) l.D.

.^
^
02 MONITOR
1
FLOW FROM
BOILER
• •

(
3.
(12 ft
1
i
1.
(=3 ft
<
CROSS-!
/v
M
10-cm (4-
SAMPLE PORT
,
7 m
3 in.)
4.
(16
m
. 9 in.)
r
SECTION
\
}
in.) i.d.
S (90° O.C.)
9 m
ft.)
1

Figure 15.  Boiler 5 exit stack (no scale).
             66

-------
CTl
                  CALIBRATION
                     GAS
                                     1/4-in. TEFLON
                         ANALYZERS
                                                                CALIBRATION GAS LINE
 S.S.     1/4-in.
3-WAY    S.S PROBE
VALVE        ,
                                                         =31 m    SAMPLE  LINE  -  1/4-in. TEFLON
                                                        (100 ft) ,
                                                                                             S.S.  CONDENSER
                                                                                               (ICE BATH)
                                              TRAILER


                                              TRANSPORT SAMPLE
                                                    PUMP
                                                                                                       STACK WALL
                                                                                                                    . PARTICIPATE
                                                                                                                       FILTER
                                                                                                                     FLOW
                                               GLASS WOOL
                                              TEFLON PUMP
                                                                                                         BUBBLE METER
                                                   '.AMPLE MANIFOLD - 1/4-in.  S.S.
                                                                                                                     O
                                                                                                                           EXHAUST
                                                               Figure  16.   CEM  system  layout.

-------
     A single Teflon sample line was used to transport the gas sample to the
NOX, CO, C02> and (^ monitors.  The gas conditioning system consisted of an
in-stack glass-fiber filter (Reeve Angel 934AH) to remove particulate matter,
followed by an ice-bath condenser to remove moisture.    The conditioned sample
gas then passed through a flow regulator, a pump, an additional  moisture
knockout jar, and into a Teflon manifold with an individual tee  connection for
each monitor.  Flow at the outlet of the manifold was  monitored  to ensure that
the sample pump was supplying a constant excess of sample gas.
     System leak checks and checks for zero drift, span drift, and response
time were performed daily on each monitor.  Guidelines set forth in 40 CFR 60,
Appendix B, Performance Specification Tests 2 and 3, were followed during this
test series.
     A three-point calibration check was performed on  each monitor at the
beginning and end of each test day.  This check covered the low, mid, and high
values of the specific pollutant concentrations measured.  Single-point cali-
bration checks were conducted between test blocks to ensure proper monitor
response.  Calibration gases were delivered through the gas sampling system
(condenser and sample line) as a check on total sample system integrity.
     Upon completion of system checks and calibration  of monitors, the sample
probe was inserted in the stack at the designated sample point.   Stack gases
were purged through the sampling system for 10 minutes, or until stable read-
ings were achieved on the monitors.  Data were then recorded for each desig-
nated test period.  The particulate filters and condenser were cleaned as
necessary between test blocks.  At the end of each test block, all monitors
were zeroed, calibrated (single or multi-point as time allowed), and prepared
for the next test block.
     All CEM's used for this test series have linear response curves.  The
three-point calibration conducted at the beginning and end of each day was
used to verify instrument linearity.  Each calibration response  had a chart
division reading and a corresponding calibration gas concentration (parts per
million, percentage).  A linear regression analysis was conducted to determine
the relationship between response and concentration or the degree of correla-
tion or linearity.
                                      68

-------
     The final reduction of data was accomplished by taking  an average  chart
reading for every 5-minute period and determining concentration by the  linear
regression equation established from the monitor calibrations.

4.3  PARTICIPATE TEST METHODS AND ANALYTICAL PROCEDURES
     Particulate was measured concurrently with the CEM data acquisition  while
Boiler 5 was firing both residual and distillate oil.
     Two individual sample trains were used to traverse the  cross-sectional
area of the stack.  One train was a standard EPA Method 5* sampling system
with a sampling temperature of 121°C (250°F), and the other  was an EPA  Method
5B* sampling system with a sampling temperature of 160°C (320°F).   Each train
consisted of a stainless steel sampling nozzle, a heated glass-lined probe, a
heated glass-fiber filter, and a series of Greenburg-Smith impingers followed
by a vacuum line, vacuum gauge, leak-free vacuum pump, dry gas meter, thermom-
eters, and a calibrated orifice.  For each train, the probe  and filter  temper-
atures were set at a predetermined temperature and monitored by the use of
multiterminal digital indicators with thermocouple leads located in each  probe
and immediately behind the Method 5 filter frits.
     The nozzle, probe, and front filter holder portions of  each sampling
train were acetone-rinsed at the end of each test.  For the  Method 5 samples,
the acetone rinse and particulate caught on the filter media were dried at
room temperature, desiccated to a constant weight, and weighed on an analyti-
cal balance.  For the Method 5B samples, the acetone rinse was evaporated at
room temperature.  The resulting rinse residue and the filter were then heated
in an oven for 6 hours at 160°C (320°F), cooled in a desiccator, and weighed
on an analytical balance.  Total filterable particulate matter was determined
by adding the rinse and filter values.

4.4  MANUAL TEST METHOD FOR NOX
     Flue gas samples were collected from each stack during  the test program
and analyzed for NO  according to procedures described in EPA Reference Method
                   A
 40 CFR 60, Appendix A, Reference Methods 5 and 5B, July 1984.
                                      69

-------
7.*  These data were used to verify the NO  CEM relative accuracy and to
                                          /\

provide additional quality assurance data for the NO  CEM system.  Most of
                                                    /\

these samples were recovered and analyzed on site.   Aliquots of the sample


solutions were also retained and analyzed in our Cincinnati  laboratory as a


check of the field analysis data.
 40 CFR 60, Appendix A, Reference Method 7, July 1984.
                                     70

-------
                                   SECTION 5
                       PROCESS DESCRIPTION AND OPERATION

     This section presents a brief process description  of  the  boiler  test
units.  Included are characterizations of the boiler and burner  and discus-
sions of the plant operating history and control  procedures.

5.1  BOILER DESCRIPTION
     Boiler Units 4 and 5 were manufactured by Babcock  and Wilcox  (B&W)  in
1974 and 1975, respectively.  Both are packed watertube boilers, Model
FM117-97 with "D"-type furnace construction (the  upper  steam drum, lower drum,
and furnace wall water tubes are in the shape of  a "D").   The  boilers are
essentially identical, but they are right- and left-hand versions  of  the same
design.  Each boiler is rated at 150,000 Ib/h steam generating capacity, or
170 x 10  to 180 x 10  Btu/h heat input (based on firing oil or  gas,  with an
economizer).  The outer dimensions for this size  boiler are 15 ft  height x 12
ft width x 29 ft length (to the nearest foot). Figure  17  shows  the layout of
Unit 5.
     Combustion air passes through a forced draft fan located  on the  boiler
house roof.  It flows down through the air preheater, which is a series  of
four steam tube coils, where it can be heated to  150° to 200°F.  Then the air
enters the windbox, which houses the Coen low excess air  (LEA) burner (which
is described more thoroughly in the next subsection).
     The flame exiting the burner travels down an open  channel within the
furnace.  The furnace sidewalls, roofs, and floor all contain  water tubes to
cool the walls and adsorb radiant heat energy from the  flame.  The rear  wall
of the furnace is a heavy duty bank of water tubes as opposed  to a refractory-
lined surface.
     The hot gases turn the corner at the back of the furnace  and  come forward
through the convection section.  This section is  where  the two drums  and

                                     71

-------
              FORCED DRAFT
                 FAN
     ROOF LINE
                                                                      SWUNG
                                                                       PORTS
   AIR
PREHEATER'
                                                                       x C0/C02 MONITOR
                                                                        • ROOF LINE -
                                                                      STACK
ECONOMIZER
                                                              BOILER GAS OUTLET
                   LOWER
                         Figure 17.   Unit  5  layout.
                                   72

-------
additional water tubes are located.   The flue gas  exits  through  the  side  of
the boiler at the front, flows up through some ductwork, passes  downward
through the economizer, and is ducted to the stack.
     Boiler feed water under pressure is preheated from  245°F  in the econo-
mizer to s300°F by the hot flue gases.  It exits the economizer, enters the
upper drum, and circulates through the water tubes of the convection and  radi-
ant sections.  The steam produced is collected in  the upper steam drum, exits
the top of the drum, and passes to the plant's header (pressure  control)
system.
     The current operating pressure of the boilers is 160 psig of saturated
steam (original design 185 psig) and the maximum pressure is 250 psig.  There
are steam vents to act as a pressure relief system if the pressure gets too
high.
     The boiler convection section has 7095 ft2 heat transfer surface area.
The water tube in the furnace contains 1072 ft2 of radiant heat  transfer  sur-
face area exposed to flame, giving a total commercial heating  surface of  8167
ft2.  The furnace volume is 1674 ft3—corresponding to approximate furnace
dimensions of 11 ft height x 6 ft width x 25 ft length (to the nearest foot).
At full load, the furnace heat release rate is approximately 165,000 Btu/h ft2
or about 100,000 Btu/h ft3.
     Each boiler is equipped with an economizer.  The economizer in Unit  4 has
2988 ft2 of bare tube surface area, but the steel  tubes  in the economizer in
Unit 5 have cast iron fins.  These fins provide additional heat  transfer
surface area and therefore Unit 5 operates cooler.  As Unit 5 is more amenable
to residual oil firing, the plant prefers to fire oil only in Unit 5.  This  is
why there are only gas tests on Unit 4, plus the fact that all the controls
for firing oil are already set for oil firing on Unit 5.

5.2  BURNER DESCRIPTION
     Each boiler contains a single burner manufactured by Coen Company,  Inc.
The burners in Units 4 and 5 are identical.  They are parallel flow type  burn-
ers capable of operating at very low excess air levels in the 5  percent excess
combustion air range, corresponding to about 1 percent flue gas  0?.   The
combustion air is accelerated to velocities sufficient to give good mixing and
                                      73

-------
complete combustion even at these air/fuel  ratios.   This burner is shown  in
Figure 18.
     The burners have 30 inch diameter throats.   These LEA burners were the
first ones of this size that Coen installed.   They  are designed to fire natu-
ral gas or residual oil (a combination of these  two fuels could be fired  in
other installations).  For natural  gas firing, the  gas is introduced in the
burner through eight gas spuds on a ring header.
     For No. 6 oil firing, the fuel is first  preheated to lower the viscosity
to 250 SSU.  It is then atomized using 160  psig  saturated steam and is  dis-
charged through the burner gun assembly running  axially down the burner
throat.  Combustion air also enters the burner throat axially (i.e., in paral-
lel flow) through two narrow bellmouth assemblies.   By adjusting these  inner
and outer concentric bellmouths, the relative air flows to three radial burn-
er zones (core, annulus, and sheath) can be varied.  Using these adjustments,
the same burner can fit an appropriate flame  pattern into similar sized fur-
naces of different shape.

5.3  OPERATING HISTORY
     In the early 1970's the plant's process  steam  demand increased due to
various process expansions.  Units  4 and 5  were  the last two, of a total  of
five, packaged boilers installed at this site.  This brought the total  capac-
ity of the steam plant to over 500,000 pounds of steam per hour.
     Since that time with increasing fuel costs,  the manufacturing plant  has
invested in heat recovery equipment.  This  has lowered the total process  steam
demand to approximately 100,000 Ib/h today.  At  this load, the steam plant
normally operates two boilers, Units 3 and  5.  Unit 5 normally swings load to
follow the process steam demand.  Unit 4 serves  as  a backup to Unit 5 and can
be fired with natural gas during times of high steam demand, such as shutdown
of major heat recovery equipment or a general plant startup.
     Originally, Units 4 and 5 were equipped  with oil-only burners supplied  by
Babcock and Wilcox.  As an additional fuel  cost  saving investment, the  plant
installed a Coen LEA burner in Unit 5 in January 1981.  During startup, this
burner had a resonant vibration problem firing natural gas; the burner  was
                                      74

-------
Figure 18.   Coen parallel  flow type LEA burner.
                      75

-------
removed and redesigned.  The redesigned burner was installed and became opera-
tional in Unit 5 in September 1982.  At the same time, an identical  burner (of
the new design) was retrofitted in Unit 4.

5.4  CONTROL PROCEDURES
     Normally the boiler plant operates in the automatic control mode.   Steam
flow and pressure measurements for the total process steam line to the  manu-
facturing plant are input signals to the master control  block of the computer
control system.  This total demand is compared to the sum of each operating
boiler's steam production.  The computer control blocks  for the fuel control
and air flow of the swing boiler are cascaded onto the master control block
and respond accordingly.  The control blocks for the combustion air and fuel
valves automatically follow the action of the opposite control  valve to keep
the air/fuel rates within limits.  The boiler feed water flow controller
follows steam output and is also subject to a drum level control override, as
is the combustion air flow to flue gas 0^ with CO set point adjustment.
     During the emissions testing of Units 4 and 5, steady-state conditions
with regard to load and excess air were required.  Because of this,  whichever
unit was being tested was operated manually and the other unit was kept on-
line to handle the swings in process steam demand.  Fuel, air, and boiler feed
water flow rates were wet manually, with the drum level  alarm set at a  pre-
determined safe level.
     All of the continuous tests for which emissions data are presented in
this report were performed at essentially steady-state conditions.  When a
problem did occur, as in the No. 2 oil tests, testing was ceased and then
recommenced when the boiler upset had passed.

5.5  PROCESS CONDITIONS DURING TESTS
     Table 34 summarizes the pertinent process data recorded by the plant
during the test runs.  Process parameters included here are steam production,
economizer inlet temperature, stack .temperature, and stack 02 concentration.
Plant values for 0? concentration differ from the data presented in Tables 2
and 3 because of the differences in monitors and sampling locations.  However,
these differences between the plant's and PEI's 02 data are consistent.

                                     76

-------
                                      TABLE 34.  BOILER PROCESS DATA
Test
block
NG-5
NG-4
NG-1
NG-3
NG-2
NG-6
NG-9
NG-10
NG-8
NG-7
5
4
7
8
9
2
3
1
6
Boiler
unit
4
4
4
4
4
4
5
5
5
5
5
5
5
5
5
5
5
5
5
Fuel
Gas
Gas
Gas
Gas
Gas
Gas
Gas
Gas
Gas
Gas
No. 2 Oil
No. 2 Oil
No. 6 Oil
No. 6 Oil
No. 6 Oil
No. 6 Oil
No. 6 Oil
No. 6 Oil
No. 6 Oil
Flue aas
02,a %
0.8
3.1
0.9
3.1
0.8
3.6
1.1
1.1
3.4
3.7
2.1
3.6
1.1
1.4
3.9
1.0
1.0
3.4
1.2
Special
conditions
_
_
-
-
-
-
_
Air preheat
-
-
Air preheat
Air preheat
_
-
-
_
Air preheat
-
Low viscosity
Steam flow,
1000 Ib/h
138
136
110
108
74
71
107
107
104
73
72
73
132
105
103
76
77
75
72
Econo-
mizer inlet
tempera-
ture, °F
683
695r
N/AC
640
N/A
560
686
687
706
608
637
654
N/A
N/A
769
641
611
N/A
629
Stack
tempera-
ture, °F
374
386
N/A
353
N/A
305
340
335
349
299
300
309
403
355
387
301
292
N/A
294
Stack
02,a %
0.5
3.1
0.8
3.0
0.5
3.5
0.7
0.5
2.7
2.9
1.5
3.1
0.8
0.9
3.1
0.8
0.8
3.2
0.9
Flue gas 02 denotes PEI monitor data and stack 02 denotes plant monitor data.

After the test program was completed, the plant calibrated the Unit  5  steam  flow meter  (orifice  plate)
with a more accurate turbine flow meter.  Based on this check and  a  review of  some  steam/header/feed
water flow rate data, we suspect the plant steam flow measurements are slightly low - ranging  from
about 5000 Ib/h at i load up to 10-15,000 Ib/h at full  load.

N/A means not available.

-------
                                  REFERENCES
 1.  Mitchell, W. J., and M. R. Midgett.   A Means to Evaluate the Performance
     of Stationary Source Test Methods.   Environmental  Science and Technology,
     10:85-88, 1976.

 2.  Oldaker, G.  B.  Condensible Particulate and Its Impacts on Particulate
     Measurements.  Draft Report.  Prepared under EPA Contract No. 68-01-4148,
     Task No. 69.  May 1980.

 3.  Peters, E. T., and J. W. Adams.  Sulfur Dioxide Interaction With Filters
     Used for Method 5 Stack Sampling.   In:  Workshop Proceedings on Primary
     Sulfate Emissions From Combustion  Sources, Volume  I - Measurement Tech-
     nology.  EPA-600/9/78-020a, 1978.   pp. 199-202.

 4.  Gushing, K.  W.  Particulate Sampling in Process Streams in the Presence
     of Sulfur Dioxide.  In:  Workshop  Proceedings on Primary Sulfate Emis-
     sions From Combustion Sources, Volume I - Measurement Technology.  EPA-
     600/9-78-020a, 1978.  pp. 202-227.

5.    PEDCo Environmental, Inc.  Comparative Evaluation  of EPA Methods 5 and
     17.  Draft Report.  Prepared under EPA Contract No. 68-02-3431, Task Nos.
     88, 103, and 163.  February 1983.

6.    PEDCo Environmental, Inc.  Method  Development and  Testing for FCCU Regen-
     erators.  Final Reports.  Prepared under EPA Contact No. 68-02-3546, Task
     Nos. 14 and 20.  February 1984.

7.    PEDCo Environmental, Inc.  Method  Development and  Testing for Boilers.
     Draft Report.  Prepared under EPA  Contract No. 68-02-3546, Task No.  15,
     June 1983.
                                    78

-------