United States Environmental Protection Agency Office of Air Quality Planning and Standards Research Triangle Park NC 27711 EMB Report 85-CHM-9 May 1985 Air Chromium Screening Study Test Report Municipal Incinerator Tuscaloosa, Alabama ------- EMISSION TEST REPORT METHOD DEVELOPMENT AND TESTING FOR CHROMIUM Municipal Refuse Incinerator Tuscaloosa Energy Recovery Tuscaloosa, Alabama ESED Project No. 85/2 EMB No. 85-CHM-9 by PEI Associates, Inc. 11499 Chester Road P.O. Box 46100 Cincinnati, Ohio 45246-0100 Contract No. 68-02-3849 Work Assignment Nos. 14, 18, and 20 PN 3615-14, 3615-18, 3615-20 Task Manager Mr. Dennis Holzschuh Emission Standards and Engineering Division U.S. ENVIRONMENTAL PROTECTION AGENCY EMISSION MEASUREMENT BRANCH RESEARCH TRIANGLE PARK, NORTH CAROLINA 27711 January 1986 ------- CONTENTS Figures iv Tables v Quality Assurance Element Finder vi Acknowledgment vii 1. Introduction 1-1 2. Summary and Discussion of Test Results 2-1 2.1 Test protocol 2-1 2.2 Particulate, hexavalent chromium, and arsenic test results 2-3 2.3 Particle size distribution test results 2-10 2.4 Process sample analytical results 2-16 • 2.5 Visible emission observations 2-17 2.6 Nitrogen oxide test results 2-17 2.7 Hi-volume sample 2-18 2.8 Total Chromium Test Results 2-20 3. Quality Assurance 3-1 4. Sampling Locations and Test Procedures 4-1 4.1 Sampling locations 4-1 4.2 Particulate, hexavalent chromium, and total chromium sample extraction and analysis 4-1 4.3 Particle size distribution 4-7 4.4 Process samples 4-8 4.5 Arsenic 4-9 5. Process Operation 5-1 5.1 Process description 5-1 5.2 Air pollution control system 5-3 5.3 Process conditions during testing 5-3 5.4 Summary of process operations during testing 5-13 (continued) ------- CONTENTS (continued) Appendices A Computer Printouts and Example Calculations A-l B Field Data Sheets B-l C Laboratory Data C-l D Sampling and Analytical Procedures D-l E Calibration Procedures and Results E-l F Quality Assurance Summary F-l G Project Participants and Sampling Log G-l H Draft Test Method for Hexavalent Chromium Emissions From Stationary Sources H-l I Draft Protocol for Determination of Total Chromium Emissions From Stationary Sources 1-1 0 Process Data 0-1 IV ------- FIGURES Number Page 2-1 Particle Size Distribution Tests PSI-2, -3, and -4 at the ESP Inlet 2-12 2-2 Particle Size Distribution Tests PSO-1, -2, and -3 at the ESP Outlet 2-13 2-3 Hi-Volume Sampler 2-19 4-1 ESP Inlet Sampling Location 4-2 4-2 ESP Outlet Sampling Location 4-3 5-1 Cross Section of Typical Modular Incinerator 5-2 ------- TABLES Number Page 2-1 Sample and Analytical Parameters 2-2 2-2 Summary of Sample and Flue Gas Data for Particulate/Cr and Arsenic Tests 2-5 2-3 Summary of Particulate and Hexavalent Chromium Emissions Data 2-6 2-4 Summary of Arsenic Emissions Data 2-10 2-5 Comparison of Particulate Concentrations as Measured by EPA Method 5 Versus Particle Size Distribution Impactors 2-14 2-6 Process Sample Analytical Results 2-16 2-7 Nitrogen Oxide Emissions Data 2-17 2-8 Summary of Total Chromium Emission Data 2-20 3-1 Field Equipment Calibration 3-3 3-2 Example Filter and Reagent Blank Analysis for Particulate 3-4 3-3 Linear Regression Data Spectrophotometer Calibration 3-4 3-4 QC Data for Total Chromium Analysis by NAA 3-7 5-1 Tuscaloosa Energy Recovery Incinerator ESP Design Specifications 5-4 5-2 Summary of Tests Conducted on the Tuscaloosa Energy Recovery Incinerator, Tuscaloosa, Alabama 5-5 5-3 Process Data for Run l--Tuscaloosa Energy Recovery May 21, 1985 5-7 5-4 Process Data for Run 2--Tuscaloosa Energy Recovery May 22, 1985 . 5-9 5-5 Process Data for Run 3--Tuscaloosa Energy Recover May 23, 1985 5-11 VI ------- QUALITY ASSURANCE ELEMENT FINDER (1) Title page (2) Table of contents (3) Project description (4) QA objective for measurement of data in terms of precision, accuracy, completeness, repre- sentativeness, and comparability (5) Sampling procedures (6) Sample custody (7) Calibration procedures and frequency (8) Analytical procedures (9) Data reduction, validation, and reporting (10) Internal quality control checks and frequency (11) Performance and system audits and frequency (12) Preventive maintenance procedures and schedules (13) Specific routine procedures used to assess data precision, accuracy, and completeness of specific measurement parameters involved (14) Corrective action (15) Quality assurance reports to management Location Section Page m 1-1 Appendix F Section 3 Appendix D Section 4 Appendix C Appendix E Section 3 Appendix D Section 4 Appendix A Appendix F Section 3 Appendix F Section 3 Appendix F Section 3 Appendix F F-2 D-l C-l E-l D-l A-l F-3 F-5 F-3 F-12 Appendix F F-4 Appendix F F-ll Appendix F F-12 VII ------- ACKNOWLEDGMENT This test program was conducted for the Emission Standards and Engineer- ing Division of the EPA Office of Air Quality Planning and Standards. Mr. Dennis Holzschuh, Emission Measurement Branch (EMB) Task Manager, provided overall project coordination and guidance and observed the test program. Mr. Ron Myers, Industrial Studies Branch (ISB) project engineer and Mr. Steve Schliesser, representing Midwest Research Institute (MRI) (an EPA contractor), monitored process operation throughout the test period. Mr. Charles Bruffey was the PEI Project Manager. Principal authors were Messrs. Charles Bruffey, David Osterhout, and Thomas Wagner. vm ------- SECTION 1 INTRODUCTION The U.S. Environmental Protection Agency (EPA) is currently evaluating several potentially toxic metals and their compounds. One of these toxic metals is chromium. Neither New Source Performance Standards (NSPS) for stationary sources nor National Emissions Standards for Hazardous Air Pollu- tants (NESHAPS) currently include chromium emissions. Available data on the emission of chromium and its impact on air quality are limited. The Emission Measurement Branch (EMB) of EPA's Environmental Standards and Engineering Division (ESED) requires contractor assistance in obtaining representative chromium emissions data from several source categories so that an accurate assessment of the potential problems can be made and appropriate regulatory action developed. PEI Associates, Inc. (under contract to ESED-EMB) performed a series of atmospheric emission tests on the Consumat Municipal Refuse Incinerators operated by Tuscaloosa Energy Recovery in Tuscaloosa, Alabama. All testing took place during the period of May 21 through 23, 1985. Triplicate tests were conducted simultaneously at the inlet and outlet of an electrostatic precipitator (ESP) used to control particulate emissions from the combined gas streams of four modular incinerators to determine the concentrations and mass emission rates of particulate matter, hexavalent chromium (Cr ), and total chromium (Cr). In addition, particle size dis- * tribution tests were conducted during the particulate/chromium tests at each 1-1 ------- location, and process samples (ESP hopper ash) were collected during each test day and analyzed for Cr and total chromium. EPA representatives obtained opacity (visible emission) data during each particulate/chromium test and also collected samples for determination of nitrogen oxides (NO ) A concentration. Prior to the commencement of the particulate/chromium tests, a single test was conducted simultaneously at each location to determine the concen- tration of inorganic arsenic. Section 2 summarizes and discusses the test results; Section 3 addresses quality assurance considerations specific to this project; Section 4 describes the sampling locations and test procedures; and Section 5 describes source operation. Appendix A presents sample calculations and computer printouts; Appendices B and C contain the field data sheets and laboratory analytical results, respectively; Appendix D details the sampling and analyt- ical procedures; Appendix E summarizes equipment calibration procedures and results; Appendix F presents a project quality assurance summary; Appendix G contains a list of project participants and a sampling log; Appendix H describes the draft test method for hexavalent chromium from stationary sources; and Appendix I presents the draft protocol for determining total chromium emissions from stationary sources. It should be noted that the U.S. EPA performed the total chromium anal- ysis of selected samples by neutron activation analysis (NAA). These data are included in Section 2 of this report. 1-2 ------- SECTION 2 SUMMARY AND DISCUSSION OF TEST RESULTS This section details the results of the sampling program. Subsections are used to identify results from each test type (i.e., particulate/Cr , particle size distribution, etc.), and results are expressed in both metric and English units where applicable. 2.1 TEST PROTOCOL Table 2-1 presents the sampling and analytical protocol followed throughout this project, the test identification, and the sampling times for each specific test type. In summary, EPA Method 5* sampling trains were used for simultaneous extraction of samples from the ESP inlet and outlet test locations. Samples were collected over a 5-hour period by isokinetic, cross-sectional traverse techniques. A total of six samples (three inlet and three outlet) were collected for determination of particulate, Cr , and total Cr concentrations. Method 5 analytical procedures were followed for the particulate analysis, and proce- dures recently developed by EPA for determination of Cr content in source emission samples were used for the Cr analysis. These latter procedures 40 CFR 60, Appendix A, Reference Method 5, July 1984. 2-1 ------- TABLE 2-1. SAMPLE AND ANALYTICAL PARAMETERS Run No. PCI-1 PCO-1 PCI-2 PCO-2 PCI-3 PCO-3 PSO-1 PSI-2 PSO-2 PSI-3 PSO-3 PSI-4 AI-1 AO-1 PCI(l-3) PCO(l-3) AI-1 AO-1 Date (1985) and time (24-h) 5/21 - 1247-1823 5/21 - 1246-1735 5/22 - 1254-1602 5/22 - 0910-1520 5/23 - 0841-1427 5/23 - 0846-1140 5/21 - 1530-1730 5/22 - 1238-1438 5/22 - 1208-1508 5/23 - 1158-1359 5/23 - 1040-1340 5/23 - 1512-1642 5/21 - 0836-1127 5/21 - 0836-1128 5/21-23 Test or sample type Participate Cr 6 Total Cr Particle size distribution Arsenic Process samples ESP hopper ash Sampling location Inlet Outlet Inlet Outlet Inlet Outlet Outlet Inlet Outlet Inlet Outlet Inlet Inlet Outlet Sample parameters Particulate Methods 1-5 X X X X X X _ - - _ - - . - - Particle size dis- tribution - - - _ - X X X X X X _ - - Cr+6 X X X X X X _ - - _ - - _ - - Total Cr X X X X X X _ _ - _ - - _ - - Arsenic Method 108 - - - _ - _ _ - _ - - X X . Analytical parameters Partic- ulate Method 5 X X X X X X _ _ - _ - - _ - _ Particle size dis- tribution - _ - _ - X X X X X X _ - _ Cr+6 X X - X - _ . - _ - - _ - X Total Cr X X X X X X . _ - _ X X _ - X Arsenic Method 108 - - - _ - _ . - . - - X X _ ro ro ------- entail extraction of the sample fractions (probe residue and filter particu- late) with an alkaline solution followed by the diphenylcarbazide colorimet- ric method.* Samples of ESP hopper ash collected during each day of testing were also analyzed for Cr using this same basic analytical technique. Particle size distribution measurements were made at each site during the particulate/Cr tests with an Andersen Mark III in-stack impactor. Four samples were collected at the ESP inlet and three were collected at the ESP outlet. Particle size fractions were analyzed gravimetrically, and size distribution curves were developed for each site. Prior to the particulate/Cr and particle size tests, a single test was conducted simultaneously at each location according to procedures described in EPA Reference Method 108.** Method 108 provides inorganic arsenic concen- tration. This test was conducted for 2 hours at each location by isokinetic, cross-sectional traverse techniques. Total arsenic content was then deter- mined by atomic absorption (AA) analysis. As indicated in Table 2-1, the selected emission and process samples were analyzed for total chromium (Cr). This analysis was performed by the U.S. EPA using NAA. The following subsections detail the results of the sampling program. 2.2 PARTICIPATE, HEXAVALENT CHROMIUM, AND ARSENIC TEST RESULTS Simultaneous Method 5* tests were conducted at the ESP inlet and outlet test locations. Samples from both sites were analyzed for particulate and inlet samples were analyzed for Cr concentrations, and the resulting data *Test Methods for Evaluating Solid Waste. U.S. EPA SW-846, 2nd ed., July 1982. ** 40 CFR 61, Appendix B, Reference Method 108, July 1984. 2-3 ------- were used to characterize the removal efficiency across the baghouse. In addition, a single test was conducted simultaneously at each location (by EPA Method 108 sampling and analytical procedures) to characterize uncontrolled and controlled arsenic emissions from this type of source. Table 2-2 summarizes pertinent sample and flue gas data for the particu- late/Cr and arsenic tests, and Table 2-3 presents the reported particulate and Cr emission results. Volumetric flow rates are expressed in cubic meters per minute (m3/min) and actual cubic feet per minute (acfm) at stack conditions. Flow rates corrected to standard conditions [20°C and 760 mmHg (68°F and 29.92 in.Hg) and zero percent moisture] are expressed as dry normal cubic meters per minute (dNm3/min) and dry standard cubic feet per minute (dscfm). Filterable particulate concentrations are expressed in milligrams per dry normal cubic meter (mg/dNm3) and grains per dry standard cubic foot (gr/dscf). Filterable particulate represents that material collected in the sample probe and on the sample filter, which were both maintained at approxi- mately 121°C (250°F). Hexavalent chromium concentrations are expressed in micrograms per gram (yg/g) and micrograms per dry normal cubic meter (yg/dNm3), where applicable. Mass emission rates are reported in kilograms per hour and pounds per hour. As reported in Table 2-2, sample volumes were fairly consistent, ranging from 1.65 to 2.41 dNm3 for the inlet trains and from 4.14 to 5.82 dNm3 for the outlet trains. Sampling times varied from the desired time of 300 min- utes for a variety of reasons. Test Runs PCI-1 and PCO-1 were abbreviated to 243 and 226.6 minutes, respectively, due to adverse weather conditions. As a result, complete cross-sectional traverses were not completed at either location. Three of twenty total points at the inlet and four of twenty 2-4 ------- TABLE 2-2. SUMMARY OF SAMPLE AND FLUE GAS DATA FOR PARTICULATE/Cr+6 AND ARSENIC TESTS Run No. PCI-1 PCO-1 PCI-2A PCO-2 PC I -3 PCO-3 PCI (In PCO (On AI-1 (Arseni AO-1 Date (1985) 5/21 5/21 5/22 5/22 5/23 5/23 let) a\ tlet) < 5/21 c) 5/21 i Sampl ing duration, min 243 226.6 150 300 300 150 erage verage 120 120 Sample volume dNnv* 1.654 4.141 2.204 5.822 2.413 4.750 2.090 4.904 1.728 2.233 dscf 58.409 146.217 77.840 205.594 85.217 167.715 73.822 173.175 61.013 78.856 Isokinetic sampling rate, * 104.9 96.4 99.2 97.0 100.4 96.3 101.5 96.3 103.7 97.8 Volumetric flow rate8 Actual mVmin 2152 2685 2498 2772 2557 3050 2402 2834 2300 2642 acfm 76,000 94,800 88,200 97,900 90,300 107,700 84,800 100,100 81,200 93,300 Standard dNmVmin 915 1204 1065 1270 1085 1325 1022 1271 997 1209 dscfm 32,300 42,500 37,600 44,900 38,300 46,800 36,100 44,900 35,200 42,700 Temperature "C 306 279 326 287 336 313 323 293 297 273 r 583 534 619 549 637 595 613 559 567 524 Moisture content. 15.6 15.7 12.4 -d 11.6 13.2 13.2 13.4 15.4 14.9 Gas compo- sition.b % °z 9.2 10.5 10.8 11.8 10.9 11.7 10.3 11.3 9.2 10.5 C6, 9.0 7.5 8.6 6.6 8.6 7.0 8.7 7.0 9.0 7.5 Gas velocity mps 14.1 22.1 16.3 22.8 16.7 25.1 15.7 23.3 15.0 21.7 fps 46.1 72.4 53.5 74.8 54.7 82.3 51.4 76.4 49.2 71.2 Static pressure, 1n.HzO -2.4 -0.60 -2.4 -0.60 -2.4 -0.60 -2.4 -0.60 -2.4 -0.60 cn aStandard conditions: 20"C (68°F), 760 ramHg (29.94 in.Hg) and zero percent noisture. bGas composition as determined from integrated bag samples collected during each test. Analysis performed with an Orsat gas analyzer. cHeasured flue gas velocity In meters per second and feet per second. Moisture not determined; broken silica gel Impinger. Inlet value used in calculations. ------- TABLE 2-3. SUMMARY OF PARTICULATE AND HEXAVALENT CHROMIUM EMISSIONS DATA Run No. PCI-1 PCO-1 PC 1-2 PCO-2 PC I -3 PCO-3 Date (1985) 5/21 5/21 5/22 5/22 5/23 5/23 Filterable concentration3 Total filterable weight, mg 297.1 314.9 264.9 520.0 323.3 397.1 Participate mg/dNm3 179.6 76.0 120.1 89.2 133.9 83.5 gr/dscf 0.078 0.033 0.052 0.039 0.058 0.036 Cr (blank corrected) ug/g <1.7d <1.7d <1.7 Total Cr+6 1n sample, ug <0.4 <0.4 <0.4 ug/dNm3 <0.24 <0.18 <0.17 Mass emission rate Participate kg/h 9.9 5.5 7.7 6.8 8.7 6.7 Ib/h 21.7 12.1 16.9 15.0 19.2 14.7 Cr*6 kg/h Ib/h <0.013 <0.029 <0.012 <0.025 <0.011 <0.024 Collection. efficiency, % Particulate 49 3 23 C/b - - - Particulate concentration in gr/dscf corrected to 12% C02 0.104 0.053 0.073 0.071 0.081 0.062 ro 01 Standard conditions: 20°C (68°F), 760 mmHg (29.94 in.Hg) and zero percent moisture. Inlet concentration - Outlet concentration Collection efficiency: Correction factor: C Inlet concentration x 100 based on Inlet and outlet concentrations corrected to 12% CO,. L12 12C HUT where C.. = concentration corrected to 12% CO^ C = concentration measured by Method 5 %CO., = as measured by Method 3 Note: A detection limit of 1.7 ug/g was established for these samples based on an approximate particulate weight of 0.3 grams using 50 ml total volume. ------- traverse points at the outlet were not sampled. Sample results were calcu- lated using the total abbreviated test time, number of points sampled, and the metered volume. Particulate stratification, which could cause large biases in particu- late measurement as a result of an incomplete sample traverse, was not a significant factor because of the generally consistent gas flow pattern at the sampling locations and the size of the particles (see Subsection 2.3). Although the results from these runs could be suspect, the overall data comparability would suggest the particulate results are representative of source conditions. Between-run data variability is primarily a process and control equipment related phenomenon (see Section 5). Run PCI-2 was found to have an excessive leak rate at the port change and was therefore repeated, but at a higher sampling rate and for 156 min- utes. Generally, the inlet sampling rates were kept lower than those at the outlet in order to prevent the filter from plugging. A cyclone was used during Run PCI-1; however, no loose particulate was collected and it was eliminated for subsequent tests. Isokinetic sampling rates ranged between 99 and 105 percent for the inlet tests and between 96 and 97 percent for the outlet tests, all of which are within the acceptable range of 90 to 110 percent. Volumetric gas flow rates at the ESP inlet ranged from 2152 to 2557 m3/min (76,000 to 90,300 acfm) and averaged 2402 m3/min (84,800 acfm) for the three particulate/Cr tests. The average volumetric flow at standard condi- tions was 1022 dNm3/min (36,100 dscfm). Flue gas temperatures ranged from 306° to 336°C (583° to 637°F) and averaged 323°C (613°F). The moisture 2-7 ------- content of the gas stream averaged 13.2 percent, and the average oxygen ((L) and carbon dioxide (CCL) contents were 10.3 and 8.7 percent, respectively. Arsenic sample and flue gas data reported in Table 2-2 are comparable to i C data associated with the particulate/Cr tests. As shown in Table 2-3, inlet participate concentrations ranged from 120.1 to 179.6 mg/dNm3 (0.052 to 0.078 gr/dscf) and averaged 144.5 mg/dNm3 (0.063 gr/dscf). The average mass emission rate for the three tests was 8.8 kg/h (19.3 Ib/h). The inlet hexavalent chromium concentration was less than 1.7 ug/g, which was determined to be the analytical detection limit for these samples. The total quantity of Cr per sample was less than 0.4 yg. At the ESP outlet, volumetric gas flow rates ranged from 2685 to 3050 m3/min (94,800 to 107,700 acfm) and averaged 2834 m3/min (100,100 acfm). The average gas flow rate at standard conditions was 1271 dNm3/min (44,900 dscfm). Flue gas temperatures ranged from 279° to 313°C (534° to 595°F) and averaged 293°C (559°F). The moisture content of the gas stream averaged 13.4 percent, and the average 0^ and CQ^ contents were 11.3 and 7.0 percent, respectively. Outlet particulate concentrations ranged between 76.0 and 89.2 mg/dNm3 (0.033 and 0.039 gr/dscf) and averaged 82.9 mg/dNm3 (0.036 gr/dscf). The average mass emission rate for the three tests was 6.3 kg/h (13.9 Ib/h). The outlet particulate concentrations corrected to 12 percent COp averaged 0.062 gr/dscf. Based on the inlet Cr analytical results, no attempt was made to analyze the outlet samples for Cr . The particulate removal efficiency of the ESP was 49 percent for Test 1, 3 percent for Test 2, and 23 percent for Test 3 and averaged 25 percent based on the average inlet and outlet particulate concentrations corrected to 12 percent COp. 2-8 ------- Two major problems were encountered during the particulate/Cr tests. First, a thunderstorm forced the abbreviation of Run 1 after about four hours. Second, during Run 2 at the inlet, an excessive leak rate was found during the port change. The inlet run was voided and repeated. The repeat test was conducted at a higher sampling rate and for half of the scheduled time of 5 hours. An analysis of the hexavalent chromium content of the ESP hopper fly ash obtained during the April 1985 pretest survey indicated Cr levels at or below the analytical detection limit for solid samples, or 0.1 yg/g. Because the concentration of hexavalent chromium in these solid samples was extremely low, the amount of alkaline extraction solution and the final dilution volume of this extract were kept at a minimum consistent with Method 3060 from Test Methods for Evaluating Solid Waste.* This proportion is 4 ml of alkaline extraction solution per gram of solid diluted to a final volume of 10 ml. Therefore, solid samples are desirable at these levels since there is no need to cut the sample filter and blank-correct for background Cr levels. Filters and/or thimbles require larger amounts of extraction solu- tion to cover the volume of material being extracted. This physical require- ment increases the analytical detection limit, as evidenced by the 1.7-yg/g detection limit established by analysis of the ESP inlet samples. No loose particulate was collected and the inlet filters were extracted as described in the analytical method. Table 2-4 summarizes the arsenic emissions data obtained at this source. *U.S. EPA SW846, 2nd ed., July 1982. 2-9 ------- The inlet sample showed a total arsenic weight of 153.8 yg or 89.0 yg/dNm3 compared with 60.0 yg (27.1 yg/dNm3) for the outlet sample. This single test indicated an arsenic collection efficiency of more than 69.5 per- cent. TABLE 2-4. SUMMARY OF ARSENIC EMISSIONS DATA Run No. AI-1 AO-1 Concentration Date (1985) 5/21 5/21 Sampling location Inlet Outlet TotaK arsenic sample weight, yg 153.8 60.9 yg/dNm3 89.0 27.3 mg/dNm3 0.089 0.027 Arsenic collection efficiency, % £69 2.3 PARTICLE SIZE DISTRIBUTION TEST RESULTS At each site an Andersen Mark III impactor was used to measure particle size distribution during each particulate/Cr test. This in-stack impactor consists of eight glass fiber filter impaction stages followed by a glass fiber backup filter. A total of three samples were collected at a single point of average velocity and temperature in each duct. Test times ranged from 90 to 120 minutes at the inlet and 120 to 180 minutes at the outlet. An initial test was conducted at the inlet using an Andersen HGLI, but the results were voided due to insufficient particulate collection. Each particle size test was conducted according to the procedures de- scribed in the Mark III operations manuals. Isokinetic sampling rates were set initially, and constant cut-point characteristics were maintained through- out the sampling period. Specifications state that the gas flow rate through the impactor at stack conditions should be maintained between 0.3 and 0.7 acfm to avoid distortion of individual stage cut-points. This criterion was 2-10 ------- met in test each. Isokinetic sampling rates ranged from 82 to 96 percent for the inlet tests and 82 to 103 percent for the outlet tests. Cumulative size distribution data points representing the total weight of particulate matter smaller than the indicated aerodynamic particle diam- eter [in micrometers (ym)] were established for each test location. The cut- points for each test were calculated by computer programs contained in "A Computer-Based Cascade Impactor Data Reduction System"* (CIDRS) developed for EPA by Southern Research Institute (SRI). All particle size results are based on a particle density of 1 g/cm3. Data reduction for the particle size runs was performed by computer programming with moisture, molecular weight, and temperature data obtained from the particulate/Cr tests. The Mark III data reduction calculations are presented in Appendix A of this report. • Figures 2-1 and 2-2 present the size distribution plot for the inlet and outlet particle size distribution tests. Table 2-5 presents a comparison of particulate concentrations obtained from the particle size tests with those obtained by Method 5 tests. Generally, the particle size distribution results indicate that most of the particulate is less than 2.5 ym at both the inlet and outlet sites and that the size distributions at both sites are very similar. For the three inlet runs (PSI-2 through -4), the size distribution curve showed that about 89 percent by weight of the particles had a nominal diam- eter of 1.0 pm or less. The calculated average particulate concentration for these runs was 166.6 mg/dNm3 (0.073 gr/dscf) compared with a three-test Method 5 average of 144.5 mg/dNm3 (0.063 gr/dscf). This indicates about a 13 * Southern Research Institute. A Computer-Based Cascade Impactor Data Reduc- tion System. Prepared for U.S. EPA under Contract No. 68-022-131, March 1978. 2-11 ------- ro ro 10.0 PARTICLE SIZE, micrometers 100 Fiaure 2-1. Particle size distribution tests PSI-2, -3, and -4 at the ESP inlet. ------- U-n ro -» 10.0 PARTICLE SIZE, micrometers 100 Figure 2-2. Particle size distribution tests PSO-1, -2, and -3 at the ESP outlet. ------- TABLE 2-5. COMPARISON OF PARTICIPATE CONCENTRATIONS AS MEASURED BY EPA METHOD 5 VERSUS PARTICLE SIZE DISTRIBUTION IMPACTORS Run No. PCI-1 PSI-2 PCI-2 PS I -3 PCI-3 PS I -4 PSO-1 PCO-1 PSO-2 PCO-2 PSO-3 PCO-3 Test location ESP inlet ESP outlet Sample type Method 5 - Particulate Particle size - Mark III Method 5 - Particulate Particle size - Mark III Method 5 - Particulate Particle size - Mark III Particle size - Mark III Method 5 - Particulate Particle size - Mark III Method 5 - Particulate Particle size - Mark III Method 5 - Particulate Particulate concentration mg/dNm3 179.6 125.8 120.1 164.1 133.9 209.8 66.6 76.0 85.3 89.2 92.4 83.5 gr/dscf 0.078 0.055 0.052 0.072 0.058 0.092 0.029 0.033 0.037 0.039 0.040 0.036 2-14 ------- percent difference in average values between the two measurements. The percentage difference between the methods is acceptable according to the applicable criterion in the Inhalable Particulate (IP) protocol.* This protocol states that a comparison of the total mass concentrations between particle size and Method 5 sample runs should not differ from the means by more than 50 percent. The size distribution curves for the first two outlet tests (PSO-1 and -2) showed that about 87 percent by weight of the particles had a nominal diameter of 1.0 ym or less. The third outlet test varied somewhat from the first two tests. The size distribution curve for the third test (PSO-3) showed that about 76 percent by weight of the particles had a nominal diam- eter of 1.0-pm or less. The average calculated particulate concentration for these runs was 81.4 mg/dNm3 (0.035 gr/dscf) compared with a three-test Method 5 average of 82.9 mg/dNm3 (0.036 gr/dscf). This indicates about a 2 percent difference between the two measurements. Both the inlet and outlet results are considered representative of particle size distribution in the gas stream at the time of testing. The data consistency and comparability to the average Method 5 results substanti- ate this conclusion. The data are also consistent with general incinerator operation. The primary chamber of the incinerator operates in a substoichiometric condition where refuse loaded into the chamber is partially oxidized. Combustible gases and particulate generated in the primary chamber flow into the second- ary chamber where combustion is completed under excess air conditions at Procedures Manual for Inhalable Particulate Samplers Operation, prepared by Southern Research Institute for EPA, Contract No. 68-02-3118, November 1979. 2-15 ------- temperatures of approximately 1150°C (2100°F). Under normal conditions, only very fine participate and condensible gases would be expected to pass from the secondary chamber. The similarity of inlet and outlet size distribution data is a function of the ESP used to control emissions. Most of the fine particulate (<2.5 ym) passes through the control device. No attempt was made to characterize Cr size distribution due to the low particulate concentrations at both sampling locations and the nondetect- able levels of Cr in the inlet samples. 2.4 PROCESS SAMPLE ANALYTICAL RESULTS Table 2-6 summarizes Cr analytical results from the ESP hopper samples collected during each day of testing. Grab samples were collected periodi- cally during the testing period. Samples were composited into a single, representative sample for each day of testing. Weighed portions of each sample type were then extracted and analyzed for Cr by procedures similar to those used .in analyzing the emission test samples. TABLE 2-6. PROCESS SAMPLE ANALYTICAL RESULTS Sample type ESP fly ash 5/21 5/22 5/23 Labora- tory ID EL566 EL567 EL568a Particu- late weight analyzed, g 10.0079 10.0540 10.0068 10.0136 Total Cr+6, yg 2.81 1.13 1.43 1.28 Cr concen- tration, yg/g 0.28 0.11 0.14 0.13 Duplicate analyses. Note: Detection limit = 0.1 yg/g. 2-16 ------- The average total Cr content of the ESP hopper ash samples was 1.8 yg, and concentrations ranged between 0.11 and 0.28 yg/g. These levels are essentially at the analytical detection limit. 2.5 VISIBLE EMISSION OBSERVATIONS Unofficial visible emission observations were made by an EPA representa- tive during each particulate/Cr test. Opacity readings were generally less than 5 percent with brief excursions to 10 and 20 percent during the second day of testing. These excursions correlated to ESP operational problems as noted by the calculated collection efficiency of 3 percent for the second test day. Raw data sheets are contained in Appendix B of this report. 2.6 NITROGEN OXIDE TEST RESULTS • Nitrogen oxide (NO ) testing was conducted by EMB personnel during the /\ particulate/Cr testing. Sample collection and analysis procedures followed those described in EPA Reference Method 7.* These data are summarized in Table 2-7. Nitrogen oxide concentrations ranged from 90 to 217 ppm and averaged 162 ppm during the test series. TABLE 2-7. NITROGEN OXIDE EMISSIONS DATA NOX Run No. TMI-1 through -8 TMI-9, -14b and TMI-16 through -18 TMI-19 through -29 Particulate/ Cr+6 Run No. PCI-1, PCO-1 PCI-2, PCO-2 PCI-3, PCO-3 Date (1985) 5/21 5/22 5/23 Concentration, ppma High 217 178 178 Low 162 90 137 Average 188 145 154 Data as reported by U.S. EPA. 3Run No. TMI-15 was not detectable and was therefore deemed invalid. 2 40 CFR 60, Appendix A, Reference Method 7, July 1984.' 2-17 ------- 2.7 HI-VOLUME SAMPLE As instructed by EPA, PEI attempted to collect sufficiently large quan- tities of particulate at the ESP outlet so that Cr and other metals of interest could be quantified. A hi-volume ambient air sample pump and filter were used as shown in Figure 2-3. A single sample run was attempted during the last test day. However, the pump motor overheated about 3i hours into the test and sampling was terminated. No attempt was made to analyze this sample based on the analytical results for Cr from the inlet and process samples. 2.8 TOTAL CHROMIUM TEST RESULTS Table 2-8 summarizes the total Cr content of selected emission samples analyzed by NAA. Analytical data as received from EPA are included in Appen- dix C of this report along with example calculations. In summary, NAA is an analytical technique dependent on the measurement of the number and energy of gamma and X-rays emitted by the radioactive isotopes produced in the sample matrix by irradiation with thermal neutrons from a nuclear reactor. Typically, the sample matrix plus appropriate stan- dards of the element(s) of interest are irradiated for a selected time period in the neutron flux core region of a research nuclear reactor. After irradia- tion and appropriate radioactive decay, a gamma-count energy spectrum is obtained by counting the sample on a nuclear detection system. As reported in Table 2-8, inlet Method 5 samples designated PCI-1 through -3 and outlet Method 5 samples designated PCO-1 through -3 were submitted for analysis. In addition, inlet particle size Run PSI-4 and outlet particle size Run PSO-3 were submitted for analysis by individual stage cut point. Process samples (ESP hopper fly ash) were also analyzed for total Cr content. 2-18 ------- ro i MANOMETER STACK WALL ICE BATH H RECIRCULATION Figure 2-3. Hi-volume sampler. ------- TABLE 2-8. SUMMARY OF TOTAL Cr EMISSION DATA ro ro o Run No. PCI-1 PC I -2 PC I -3 PCO-1 PCO-2 PCO-3 PSI-4 PSO-3 1 2 3 Sample type and location Filterable participate ESP inlet Filterable participate ESP inlet Filterable particulate ESP inlet Filterable particulate ESP outlet Filterable particulate ESP outlet Filterable particulate ESP outlet Particle size ESP inlet Stage 0 Stage 1 Stage 2 Stage 3 Stage 4 Stage 5 Stage 6 Stage 7 Backup Particle size ESP outlet Stage 0 Stage 1 Stage 2 Stage 3 Stage 4 Stage 5 Stage 6 Stage 7 Backup ESP fly ash ESP fly ash ESP fly ash Total3 particulate collected, 9 0.2971 0.2649 0.3278 0.3149 0.5200 0.3971 0.0068 0.0001 0.0002 0.0000 0.0007 0.0024 0.0028 0.0063 0.0902 0.0139 0.0011 0.0009 0.0006 0.0009 0.0007 0.0014 0.0037 0.0567 _ - - Particulate sample weight analyzed by NAA, g 0.2971 0.2649 0.3278 0.3149 0.5200 0.3971 0.0068 0.0001 0.0002 0.0000 0.0007 0.0024 0.0028 0.0063 0.0902 0.0139 0.0011 0.0009 0.0006 0.0009 0.0007 0.0014 0.0037 0.0567 0.1039 0.1254 0.1144 Total Crc results by NAA, ug 52.1 50.0 67.8 47.1 142.9 58.1 9.95 7.79 9.36 6.72 9.61 8.59 8.28 10.56 14.12 20.29 8.62 8.98 9.26 7.39 6.85 9.60 10.67 14.19 45.7 89.5 60.6 Total Crd concen- tration by NAA, ug/g 175.4 188.8 206.8 149.6 274.8 146.3 f f f f f f f f f 801.4 f f f f f f f f 439.8 713.7 529.7 Total Cre content of sample, "9 52.1 50.0 67.8 47.1 142.9 58.1 f f f f f f f f f 11.14 f f f f f f f f _ - - Total Cr concentration, ug/dNm3 _ 31.5 22.7 28.1 11.4 24.5 12.2 f f f f f f f f f 12.88 f f f f f f f f _ - - Total Cr mass emission rate, kg/h 0.0017 0.0015 0.0018 0.0008 0.0019 0.0010 Total particulate (acetone rinse residue and filter) collected during sample run. ug Cr for filter/acetone blank). Particle-size data are uncor- Particulate weight analyzed by NAA. c Total Cr results by NAA. Run Nos. PCI and PCO are blank corrected values rected. (=8 ug Cr for filter stage blank.) Total Cr(C) divided by particulate weight analyzed by NAA(b). Total Cr concentration (ug/g) multiplied by total particulate weight collected(a). Assumed tn bo below the detection limit since sample values are less than two times the stnadard deviation nf the blank values. ------- The total Cr content of the inlet samples on a yg/g basis ranged between 175.4 and 206.8 yg/g. Total Cr concentrations on a yg/dNm basis ranged 3 between 22.7 and 31.5 yg/dNm with corresponding mass emission rates ranging between 0.0015 kg/h (0.0033 Ib/h) and 0.0018 kg/h (0.0040 Ib/h). The total Cr content of the outlet samples on a yg/g basis ranged be- tween 146.3 and 274.8 yg/g. The total Cr concentration on a yg/dNm basis 3 ranged between 11.4 and 24.5 yg/dNm with corresponding mass emission rates of 0.0008 kg/h (0.0018 Ib/h) and 0.0019 kg/h (0.0042 Ib/h), respectively. All inlet and outlet total Cr data have been corrected for a filter/acetone Cr blank level of 21 yg, which is reasonable considering the levels of Cr detected in these samples. Total Cr in the ESP fly ash samples ranged from 439.8 yg/g to 713.7 yg/g. The data were determined from samples obtained during each test. Particle size samples from each location were analyzed by individual stages in an attempt to characterize total Cr by size fraction. As reported in Table 2-8, the total Cr content of individually-loaded stages ranged from 6.7 to 14.1 yg for the inlet sample, and 6.9 to 20.3 yg for the outlet sam- ple. Filter blank Cr levels ranged from 6.3 to 14 yg and averaged about 8 yg per filter for a set of eight filters (see Section 3). Because the filter blank levels of total Cr are similar to measured Cr values on each stage and the sample values are less than two times the standard deviation of the blank values, no reasonable conclusions can be drawn regarding total Cr size distri- bution. The uncorrected data reported in Table 2-8 indicate that some Cr is present in the sample fractions of Stage 7 and the backup filter, which represent particle sizes of less than 2 ym. 2-21 ------- SECTION 3 PROJECT QUALITY ASSURANCE The application of quality assurance procedures to source emission measurements ensures accurate emission-testing results. Quality assurance guidelines provide the detailed procedures and actions necessary for defining and producing acceptable data. In this project, five documents were used in the preparation of a source-specific test plan that would ensure the collec- tion of acceptable data: the EPA Quality Assurance Handbook Volume II, EPA-600/4-77-0271; the PEI Emission Test Quality Assurance Plan; the PEI Laboratory Quality Assurance Plan; Determination of Hexavalent Chromium Emissions From Stationary Sources, December 13, 1984; and EPA Protocol for Emissions Sampling for Both Hexavalent and Total Chromium, February 22, 1985. Two of these are PEI's general guideline manuals and define the company's standard operating procedures followed by the company's emission testing and laboratory groups. In this specific test program, which was reviewed by EPA's Emission Measurement Branch, the following steps were taken to ensure that the testing and analytical procedures produced quality data: 0 A sample of the ESP hopper catch w$s obtained during the April pretest survey and analyzed for Cr 6 content. These data were used to+define sampling times and rates so that a quantifiable level of Cr 6 could be collected. 0 Calibration of all field sampling equipment. 0 Checks of train configuration and calculations. 3-1 ------- 0 Onsite quality assurance checks, such as leak checks of the sam- pling train, pitot tube, and Orsat line and onsite quality assur- ance checks of all test equipment prior to use. 0 Use of designated analytical equipment and sampling reagents. 0 Internal and external audits to ensure accuracy in sampling and analysis. Table 3-1 lists the specific sampling equipment used to perform the particulate/Cr , particle size distribution, and arsenic tests as well as the calibration guidelines and limits. In addition to the pre- and post-test calibrations, field audits were performed on the metering systems and temper- ature-measurement devices used during sampling. These data are summarized in Table 3-1, and copies of the field audit data sheets are presented in Appen- dix B of this report. The PEI project manager performed the onsite sample calculations, and computer programming was used to validate the data upon return to PEI's Cincinnati laboratory. Minor discrepancies between the hand calculations and computer printouts are due primarily to rounding off of values. Computerized example calculations are presented in Appendix A. The following subsections summarize the quality assurance activities performed during the analytical phase of this project. As a check of the gravimetric analytical procedure, blank filter and reagent (acetone) were analyzed in a fashion similar to that used for the actual field samples. Table 3-2 summarizes the blank analysis data. These data indicate good analytical technique. Emission and process samples were analyzed in two separate batches. Table 3-3 summarizes the linear regression data of the spectrophotometer calibration for the two days. 3-2 ------- TABLE 3-1. FIELD EQUIPMENT CALIBRATION Equipment Meter box Pi tot tube Digital indicator Stack thermocou- ple Orsat analyzer Impinger thermocouple Balance Barometer Dry gas thermometer Probe nozzle ID No. FB-3 FB-8 FB-11 FB-12 020 516 220 262 139 206 141 1-1 1-5 Hettler No. 743985 227 FB-3 FB-8 FB-11 FB-12 2-120 4-101 5-108 Part, size Calibrated against Wet test meter Standard pi tot tube Millivolt signals ASTM-3F Standard gas ASTM-3F Type S weights NBS traceable barometer ASTM-3F Call per Allowable error AH P iO.15 (Y ±0.5% V post-test) Cp +0.01 +0.5% ±1.5* (±2% saturated) ±0.5% ±2°F ±0.5 g ±0.10 in.Hg (0.20 post-test) ±5°F On ±0.004 In. Actual error AHP: -0.02; Y: -0.007% AHQ: -0.06; Y: -0.010% 6H0: -0.04; Y: -0.014% AH0: 0.0; Y: .0.017% - -0.4% +0.4% +0.2% +0.3% -0.4* (CO,) 0.0% (02) + 1°F + 1°F 0.0 g +0.02 In.Hg In: -2°F; Out: -2°F In: -4°F: Out: -3°F In: -1°F; Out: -4°F In: -1°F; Out: 0°F 0.000 in. 0.000 in. 0.000 in. 0.002 in. 0.002 in. 0.001 in. 0.001 in. Within allowable limits X X X X X X X X X X X X X X X X X X X X X X X X X X X Comments Y = 0.029; Audit AH9 0.08 Y = 0.019; Audit AHP 0.01 Y = 0.014; Audit AHP 0.01 Not performed (Field audit results) Visually inspected on site Visually inspected on site Field audit results: 0.2* Field audit results: 0.3* Audit value C02 and Or * 5.2% Inlet Inlet Inlet Outlet Outlet Inlet Outlet CO CO ------- TABLE 3-2. EXAMPLE FILTER AND REAGENT BLANK ANALYSIS FOR PARTICULATE Sample type and filter number Participate - 8510139 Reeve Angel 934 AH Acetone blank3 Original tare weight, mg 363.3 102,545.1 Blank weight, mg 363.8 102,547.1 Blank value 0.5 mg 0.013 mg/gb 188 ml evaporated and desiccated before weighing. 'Exceeds blank value limitations (0.01 mg/g used in calculations). TABLE 3-3. LINEAR REGRESSION DATA SPECTROPHOTOMETER CALIBRATION Sample description Process, duplicates, plus spike Inlet and blank Date (1985) 6/5 6/11 Cr+6 standard concen- tration, yg/ml 0.0 0.1 0.2 0.3 0.4 0.5 0.0 0.1 0.2 0.3 0.4 0.5 Absorb- ance 0.000 0.130 0.262 0.390 0.522 0.656 0.000 0.130 0.262 0.386 0.513 0.636 Y-Intercept -0.0008 0.0039 Slope 1.3097 1.2666 Correlation coefficient 0.99998 0.99986 3-4 ------- The analytical detection limit established for the two days was less than 0.004 yg/ml for an absorbance of 0.005 above the linear regression intercept. This detection limit corresponds to the same total micrograms of chromium VI as in previous reports. The established detection limit for the inlet particulate samples was 1.7 yg/g because of the small amount of par- ticulate collected on these filters. Because the concentrations of hexavalent chromium in these solid samples were extremely low, the amount of alkaline extraction solution and the final dilution volume of this extract were kept at a minimum consistent with Method 3060 from Test Methods for Evaluating Solid Waste.* This proportion is 4 ml of alkaline extraction solution per gram of solid diluted to a final volume of 10 ml. Filters and thimbles require larger amounts of extraction solution to cover the volume of material being extracted. This physical requirement increased the detection limit. The filter/acetone blank contained 0.87 yg of Cr . This value was used to correct the inlet particulate values. Duplicate and spike analyses were performed on the ESP fly ash sample taken on May 23, 1985. The results of the duplicate analyses were 0.14 and 0.13 pg/g, which is just above the detection limit of 0.1 ug/g for process samples having a large quantity of loose particulate available for analysis and requiring no reagent or filter blank correction. The spike recovery was 103 percent. One problem was encountered with the analysis of the inlet particulate samples. After the alkaline extract was filtered, the samples were slightly cloudy (white opaqueness). This interfered with the colorimetric determina- tion, and the absorbance of an aliquot not containing the color reagent was subtracted from the absorbance of the sample to correct for this interferent. *U.S. EPA SW846, 2nd ed., July 1982. 3-5 ------- Arsenic samples were analyzed on June 4, 1985. The least-squares fit of the data to quadratic equations for the graphite furnace atomic absorption calibration gave a correlation coefficient of 0.9990. The results of blank analyses were 9.60 yg for the filter and <0.006 mg/liter (detection limit) for the rinse and impinger samples. The filter blank value is typical for glass fiber filters. Duplicate analysis of the impinger fraction for the outlet runs gave 6.22 and 5.27 yg, respectively. Spike recoveries were 95.4, 92.1, 101.9, and 0 percent for the filter, rinse, impingers, and bomb frac- tions, respectively. Table 3-4 presents QC data relative to the total Cr analysis by NAA. Duplicate, audit, and blank data are presented. As discussed in Section 2.8, a filter/acetone Cr blank level of 21 yg is reported. Total Cr results for the inlet and outlet particulate emission samples were corrected for blank levels since the measured Cr content ranged between two and four times the blank values. Particle size Cr data were not corrected for filter blank Cr levels since measured Cr values were less than two times the blank Cr values for the majority of the stages analyzed. This seems prudent considering the low levels of total Cr observed in the samples. 3-6 ------- TABLE 3-4. QC DATA FOR TOTAL CHROMIUM BY NAA Lab No. ESP fly ash - Run 1 EL 566 ESP fly ash - Run 3 EL 568 NB5 fly ash Alkaline extract Type I H20 Method 5 filter/ acetone Particle size filter - Stage 0 Particle size filter - Stage 1 Particle size filter - Stage 2 Particle size filter - Stage 3 Particle size filter - Stage 4 Particle size filter - Stage 5 Particle size filter - Stage 6 Particle size filter - Stage 7 Particles size filter - Backup Sample type Duplicate analysis Duplicate analysis Audit Blank Blank Blank Blank Blank Blank Blank Blank Blank Blank Blank Blank Results total Cr, ug (except as noted) 45.7; 44.0 60.6; 36.6 201 yg (196 ug/g accepted) Not detected Not detected 21.0 9.40 7.78 14.09 6.67 6.75 . 5.76 6.36 8.11 9.68 3-7 ------- SECTION 4 SAMPLING LOCATIONS AND TEST PROCEDURES This section describes the sampling sites and the test methods used to characterize participate, chromium, and arsenic emissions. 4.1 SAMPLING LOCATIONS Flue gas samples were extracted simultaneously from the inlet duct and outlet stack of the ESP. Figures 4-1 and 4-2 show the sampling locations. At the ESP inlet, two 10.2-cm (4-in.) i.d. sampling ports were located approximately 5.9 duct diameters downstream and 1.7 duct diameters upstream from the nearest flow disturbances in the 1.8-m (5-ft 11-in.) i.d. duct. At the ESP outlet, two 10.2-cm (4-in.) i.d. sampling ports were located approxi- mately 5.7 stack diameters downstream and 2.1 stack diameters upstream from the nearest flow disturbances in the 1.6-m (5-ft 3i-in.) i.d. stack. Both locations conformed to the minimum requirements for sampling port locations specified in EPA Reference Method 1.* 4.2 PARTICIPATE AND HEXAVALENT CHROMIUM SAMPLE EXTRACTION AND ANALYSIS Flue gas samples were simultaneously collected at the ESP inlet and outlet test locations according to procedures outlined in EPA Reference Method 5.** 40 CFR 60, Appendix A, Reference Method 1, July 1984. 40 CFR 60, Appendix A, Reference Method 5, July 1984. 4-1 ------- *35 ft 10 ft FLOW DUCT TO ESP I ro ~ ^. t HEAT RECOVERY BOILER EXHAUST JS25 ft UNIT NO. 1 II TWO 10.2 cm (4 in.) i.d. SAMPLING PORTS CROSS-SECTION (LOOKING TOWARDS ESP) »1.8 m (5 ft 11 in.) i.d. WALL GRADE Figure 4-1. ESP inlet sampling location. ------- «3.4 m(ll ft) PORTS' /J 1 ^9.1 m ( «5.5 m \ ^ Q 1 ^^^^ 30 ft) 18 ft) ' en n l-l NO SAMPLE X LOCATION r i i I i •~~ — — — ^^^m ^^•B MH^ ^^^ ••^H ^^^m + 1 J SAMPLING PLATFORM CROSS-SECTION i S*" " """"X. / 31/4 in .Yi V 1 TWO 10.2 cm (4 in.) SAMPLING PORTS ^ FLUE GAS FROM I.D. FAN RRAOF Figure 4-2. ESP outlet sampling location. 4-3 ------- Initially, the collected samples were analyzed gravimetrically by Method 5 to determine particulate concentration and mass emission rates. At the completion of the gravimetric analysis, the samples were prepared and ana- lyzed for Cr according to procedures described in a draft EPA method en- titled "Determination of Hexavalent Chromium Emissions From Stationary Sources." A copy of the draft method is contained in Appendix H of this report. Before sampling began, velocity, static pressure, molecular weight, moisture content, and temperature were measured to define sampling rates and nozzle sizes are described in EPA Reference Methods 1 through 4.* The degree of turbulent flow at each location also was assessed according to procedures described in EPA Reference Method 2.* In this method, the face opening of the Type-S pitot tube is aligned perpendicularly to the duct cross-sectional plane, designated "0-degree reference." Null (zero) pitot readings obtained at a 0-degree reference indicate an acceptable flow condition at a given point. If the pitot reading is not zero at 0-degree reference, the pitot is rotated (up to 90 degrees ± yaw angle) until a null reading is obtained. The value of the rotation angle (yaw) is recorded for each point and averaged across the duct. Method 2 criteria stipulate that average angular rotations greater than ± 10 degrees indicate turbulent (nonaxial) flow conditions in the duct(s). This procedure was used to check several traverse points at each location. In each case, null pitot readings were observed at the 0- degree reference. These data, together with the velocity and temperature profiles established for each location, indicated acceptable flow patterns that would enable the extraction of representative samples at each site. 40 CFR 60, Appendix A, Reference Methods 1 through 4, July 1984. 4-4 ------- A total of 20 sampling points were used to traverse the cross-sectional areas of the ducts. At each site, each point was to be sampled for 15 min- utes, thus yielding a total test time of 300 minutes. However, the first test was abbreviated due to adverse weather, and the second inlet test was also abbreviated due to an excessive leak rate following the first half of the test. This second test (PSI-2) was run at a higher sampling rate for 150 minutes. The testing and analytical procedures used are described briefly here, and detailed procedures are presented in Appendix D. 4.2.1 Velocity and Gas Temperature A Type-S pitot tube and an inclined draft gauge manometer were used to measure the gas velocity pressures at the test sites. Velocity pressures were measured at each sampling point across the duct to determine an average value. Measurements were taken in the manner prescribed in EPA Reference Method 2.* The temperature at each sampling point was measured with a ther- mocouple and digital readout. 4.2.2 Molecular Weight Flue gas composition was determined in accordance with the basic proce- dures described in EPA Reference Method 3.* Grab samples were collected prior to the start of sampling to establish baseline contents of oxygen, carbon dioxide, and carbon monoxide. Integrated bag samples were collected during each test and were analyzed with an Orsat gas analyzer. The gas composition at each test site remained reasonably consistent throughout the test series. 40 CFR 60, Appendix A, Reference Methods 2 and 3, July 1984. 4-5 ------- 4.2.3 Particu1ate/Cr+6 Samples of particulate, Cr, and total Cr were collected as specified in EPA Reference Method 5.* All tests were conducted isokinetically by regulating the sample flow rate relative to the gas velocity in the duct (as measured by the pitot tube and thermocouple attached to the sample). The basic sampling train consisted of a heated glass-lined probe, a heated 7.6-cm (3-in.) diameter glass-fiber filter (Whatman Reeve Angel 934 AH), and a series of five Greenburg-Smith impingers followed by a vacuum line, vacuum gauge, leak-free vacuum pump, dry gas meter, thermometers, and a calibrated orifice. For determination of particulate concentration, the nozzle, probe, and filter holder portions were rinsed with acetone at the end of each applicable test. The acetone rinse and particulate caught on the filter media were dried at room temperature, desiccated to a constant weight, and weighed on an analytical balance. Total filterable particulate matter was determined by adding these two values. Upon completion of the gravimetric analysis, the inlet sample fractions were prepared and analyzed for Cr according to procedures recently devel- oped by EPA. In summary, the samples were digested in an alkaline solution and analyzed by the diphenylcarbazide colorimetric method.** Selected samples were then shipped to EPA, where total chromium content of the samples was determined by NAA. 40 CFR 60, Appendix A, Reference Method 5, July 1984. ** Test Methods for Evaluating Solid Waste. U.S. Environmental Protection Agency, SW-846, 2nd ed., July 1982. 4-6 ------- The volume of water collected in the impinger section of the sampling train(s) was measured at the end of each sample run to determine the moisture content of the flue gas. The contents of the impingers were transferred to a polyethylene container. The impingers and all connecting glassware, includ- ing the back half of the filter holder, were rinsed with distilled water and the rinse was added to the container. These samples are being held for further analysis if desired. 4.3 PARTICLE SIZE DISTRIBUTION Samples for particle-size distribution measurements were collected at the ESP inlet and outlet by an Andersen Mark III impactor. This in-stack impactor consists of eight cut-point stages and a backup filter. The impac- tor was preceded by a 15-pm cyclone precutter. The sampled gas stream enters the system through the precutter. Particles with sufficient inertia are impacted against the sides of the cyclone. Smaller particles flow with the gas stream and exit the cyclone and into the main impactor. Then, particles with sufficient inertia are impacted on the first stage filter with smaller particles passing onto the second stage, and similarly to each succeeding stage. Finally, a glass fiber filter removes all particles remaining in the gas stream downstream of the final stage. This type of impactor was used at both sites since the particulate concentrations were low at both sites. Three samples were collected at the ESP inlet and outlet from single points in the ducts that represented average velocities and temperatures. At the inlet, Tests PSI-2 and -3 were run for 120 minutes, with Test PSI-4 running for 90 minutes. At the outlet, Test PSO-1 was conducted for 120 minutes, and Tests PSO-2 and -3 were conducted for 180 minutes. Isokinetic 4-7 ------- sampling rates were set initially, and constant cut-point characteristics were maintained throughout the sampling period. At the completion of each test, the impactor samples were recovered according to procedures described in the Mark III operations manual. Each recovered fraction was subjected to a gravimetric analysis using EPA Method 5 procedures. Size distribution curves were established repre- senting the total weight percent of particulate matter smaller than the indicated aerodynamic particle diameter in micrometers. Cut-points for the eight Mark III impactor stages were calculated by computer programs contained in "A Computer-Based Cascade Impactor Data Reduc- tion System" (CIDRS) developed for EPA by Southern Research Institute (SRI).* All particle size results are based on a particle density of 1 g/cm3. Data reduction and intermediate results calculations for both types of impactors were performed by the CIDRS program, with moisture contents obtained from the particulate tests. Samples from Runs PSI-4 and PSO-3 were analyzed for total Cr by NAA. 4.4 PROCESS SAMPLES Grab samples of ESP hopper ash were collected periodically during each particulate/chromium test. The samples from each test were composited so that one representative sample per test was available for analysis. The samples were extracted and analyzed for Cr in accordance with procedures Southern Research Institute. A Computer-Based Cascade Impactor Data Reduc- tion System. Prepared for U.S. Environmental Protection Agency under Con- tract No. 68-022-131, March 1978. 4-8 ------- similar to those used for the Cr analysis of the particulate samples. An aliquot of each sample was also analyzed for total Cr. 4.5 ARSENIC Arsenic concentrations were measured by EPA Reference Method 108.* All tests were conducted isokinetically by regulating the sample flow rate to correspond to the gas velocity in the duct (as measured by the pi tot tube and thermocouple attached to the sample probe). The basic sampling train con- sisted of a heated glass-lined probe, a heated 7.6-cm (3-in.) diameter glass- fiber filter (Whatman Reeve Angel 934 AH), and a series of five Greenburg- Smith impingers followed by a vacuum line, vacuum gauge, leak-free vacuum pump, dry gas meter, thermometers, and a calibrated orifice. For determination of arsenic concentrations, the nozzle, probe, and filter holder portions were rinsed with 0.1 N NaOH at the end of each appli- cable test. The filter and solids contained in the 0.1 N NaOH rinse of the front half of the sampling train were prepped, combined, and analyzed for arsenic (by atomic absorption). The volume of water collected in the impinger section of the sampling train was measured at the end of each sampling run to determine the moisture content of the flue gas. The contents of the impingers were transferred to a polyethylene container. The impingers and all connecting glassware (includ- ing the back half of the filter holder) were rinsed with 0.1 N NaOH, and the rinse was added to the container. The contents of the impingers and 0.1 N NaOH rinse also were analyzed for arsenic by atomic absorption. *40 CFR 61, Appendix B, Reference Method 108, July 1984. 4-9 ------- SECTION 5 PROCESS OPERATION The following process description was prepared by Midwest Research Institute (MRI), the EPA New Source Performance Standards (NSPS) contractor. Personnel from MRI were on site during the test program to monitor and record all pertinent incinerator and ESP data. 5.1 PROCESS DESCRIPTION Tuscaloosa Energy Recovery incinerator facility consists of four modular municipal refuse incinerators manufactured by Consumat Systems, Incorporated, and installed in 1984. Each incinerator has a rated capacity of 80 Mg/day (90 tons/day) and typically operates 24 hours per day, 5 days a week. During the month preceding testing, each incinerator burned an average of 56 Mg/day (63 tons/day). Exhaust from the four incinerators is fed through two heat recovery boilers to produce 24,900 kg (55,000 Ib) of steam per hour. Ap- proximately 99 percent of the refuse incinerated is supplied by residential sources in eight nearby municipalities. The remaining 1 percent [4.5 Mg/day (5 tons/day)] consists of scrap tires from the nearby B.F. Goodrich plant that purchases the steam produced. Large metal items are manually sorted from the refuse prior to incineration. Figure 5-1 shows a cross section of a typical modular incinerator con- sisting of a primary and a secondary chamber. During startup, natural gas is introduced via one burner in each chamber. Refuse is loaded into the primary chamber and is partially oxidized under starved air, or substoichiometric, 5-1 ------- Figure 5-1. Cross section of typical modular incinerator. 5-2 ------- conditions. (Primary chambers of modular starved air incinerators typically operate at 25 to 50 percent below stoichiometry.) The combustible gases and particulates generated in the primary chamber flow into the secondary chamber where combustion is completed under excess air conditions, typically 50 to 100 percent above stoichiometry. Temperature in the primary chamber of each incinerator is maintained between 540° and 760°C (1000° and 1400°F). Sec- ondary chamber temperatures are typically 1150°C (2100°F). No auxiliary fuel is used during normal operation. 5.2 AIR POLLUTION CONTROL SYSTEM Particulate emissions are controlled by an electrostatic precipitator (ESP) manufactured by Precipitair Pollution Control Company of Longview, Texas. Exhaust from the four incinerators is routed through the ESP prior to exiting through a single stack. An induced draft (ID) fan is located after the ESP and before the stack. Design specifications for the ESP are pre- sented in Table 5-1. 5.3 PROCESS CONDITIONS DURING TESTING All tests were conducted while the four incinerator modules were opera- ting normally at approximately 320 tons of refuse per day (90 percent of capacity). Table 5-2 outlines the tests conducted on the incinerator. The lower chamber temperatures and upper chamber temperatures were monitored and controlled to operate in the typical ranges of 980° to 1200°F and 2080° to 2120°F, respectively. The following incinerator process parameters were recorded every 15 minutes during testing: lower chamber and upper chamber temperatures for each of the four modules (Nos. 1 through 4), steam flow, 5-3 ------- TABLE 5-1. TUSCALOOSA ENERGY RECOVERY INCINERATOR ESP DESIGN SPECIFICATIONS Collection efficiency, % 50 Inlet dust loading, gr/dscf 0.06 Outlet dust loading, gr/dscf 0.03 Gas volume, acfm 76,000 Gas temperature, °F 350 No. of fields 2 Collection plate area, ft2 10,617 Specific collection area, ft2/!,000 acfm 140 Collection plate height, ft 28 Collection plate length, ft 7.3 Aspect ratio 0.52 Gas velocity, ft/s 4.18 Gas passage spacing, 1n. 10 No. of gas passages 13 No. of plates 28 Secondary voltage, kV 45 Secondary current, mA 300 ESP power, kVA 27 5-4 ------- TABLE 5-2. SUMMARY OF TESTS CONDUCTED ON THE TUSCALOOSA ENERGY RECOVERY INCINERATOR, TUSCALOOSA, ALABAMA3 Sampling point ESP inlet ESP outlet Outside observa- Test type Particulate concentration Particle size Particulate concentration Particle size NO concentration A Visible emissions Test Method EPA Method 5 Anderson Mark III EPA Method 5 Anderson Mark III EPA Method 7 EPA Method 9 No. per run 1 1 1 1 ~4b tion point ESP dust hopper Fly ash Grab •10 Tests consisted of one run per day on May 21, 22, and 23, 1985. Method 5 samples on the first day were gathered for arsenic analyses. A116other samples to be analyzed for particulate, total chromium, chromium , and cadmium. Each observation period was 6 minutes in duration. These data were collected to support transmissometer data. 5-5 ------- steam temperature, steam pressure, ID fan amps, carbon monoxide concentra- tion, and stack gas opacity. Integrated steam flow values were recorded daily at the beginning and end of the test period. Tables 5-3, 5-4, and 5-5 present the process parameters recorded during Runs 1, 2, and 3, respec- tively. Appendix J contains charts of the opacity and steam parameters recorded during the test period. A local power outage during a thunderstorm stopped the inlet Method 5 and particle sizing sampling at 6:20 p.m. during Run 1. During the temporary power loss, the emergency venting system was engaged to allow the incinerator exhaust to bypass the ESP and stack. Sampling during Run 2 was interrupted due to circuit overload from the sampling equipment. Sampling resumed after an additional electrical circuit was used to power the sampling equipment. Air inleakage at the ID fan flange was discovered and sealed during Run 3. The ESP operation did not remain steady during the three-day test period, although it was operating as usual according to plant personnel. Prior to Run 1, plant personnel adjusted the electronic controls to obtain the maximum secondary voltages to both ESP fields and to frequent a short in the inter- nals. Plant personnel also questioned the accuracy of the voltage and cur- rent meters. During Run 1, power to the ESP inlet field was interrupted several times because the transformer-rectifier tripped off. Primary and secondary current meters both registered a zero value during Runs 1 and 2 even when the inlet field was apparently operating. The averge power level to the ESP during Run 1 was estimated to be 1.70 kVA, representing 6 percent of the design power level of 27 kVA. During Run 2, from 8:00 to 11:45 a.m., the inlet field tripped off occasionally and then stayed off after 11:45 a.m. 5-6 ------- TABLE 5-3. PROCESS DATA FOR RUN 1—TUSCALOOSA ENERGY RECOVERY May 21, 1985 en --J Time a.m. 7:30 8:00 8:15 8:30 8:36 8:45 9:00 9:15 9:30 9:45 10:00 10:15 10:30 10:45 11:00 11:15 11:27 11:30 p.m. 12:45 12:46 12:47 1:00 1:15 1:30 1:45 2:00 2:15 Unit No. 1 Unit Lower Upper Lower chamber chamber chamber temp., temp., temp., 'F *C *F 1194 1143 1210 1150 1223 1153 1195 1146 MS inlet and outlet 1166 1146 1152 1149 1131 1150 1130 1148 1104 1148 1095 1146 1043 1146 1034 1153 1014 1144 985 1 1 55 986 1145 H5 inlet and outlet 986 1148 966 1 1 50 M5 outlet started M5 inlet started 970 1146 981 1147 985 1149 994 1 1 53 1010 1150 1031 1148 1400 1253 1172 1171 test 1144 1079 1058 1054 1063 1081 1046 1036 1039 1031 1055 test 1083 1054 990 982 969 958 952 948 No. 2 Unit No. 3 Upper Lower Upper chamber chamber chamber temp., temp., temp., •C *F 'C 1140 1134 1140 1142 for arsenic 1138 1132 1142 1143 1146 1149 1142 1138 1140 1140 1141 for arsenic 1141 1144 1142 1143 1142 1140 1141 1142 1106 1137 1188 1215 started 1188 1145 1091 1090 1091 1076 1054 1054 1052 1041 1044 ended 1045 1067 1061 1052 1045 1038 1035 1030 1146 1147 1147 1150 1148 1120 1153 1145 1145 1144 1146 1138 1144 1142 1148 1127 1146 1145 1143 1144 1145 1147 1150 Unit No. 4 Lower chamber temp., •F 1178 1197 1236 1220 1218 1145 1151 1138 1075 1036 1067 1072 1123 1125 1126 1123 1069 1062 1066 1075 1089 1084 1080 Upper chamber temp., •c 1146 1140 1146 1149 1148 1165 1152 1147 1148 1146 1147 1145 1142 1145 1148 1165 1142 1144 1147 1148 1150 1150 1148 Steam temp., •F 435 435 435 435 435 430 430 430 435 435 435 435 435 435 435 435 430 430 430 430 430 430 430 Steam Steam pressure, flow, psi Ib/h 355 355 355 355 350 350 355 355 355 360 360 360 360 360 355 355 350 350 350 350 355 355 355 65.000 66,000 65,000 64,000 62,000 62,500 63,000 63,000 63,000 68,000 67,500 67,500 65,000 66,000 63.000 62,500 62,000 63.000 62,000 63,000 64,000 64,000 63,500 I.D. fan, amps 210 205 200 200 200 200 200 200 200 205 200 200 200 200 200 200 195 200 200 200 200 200 200 Opac i t} % 4 6 7 6 5 4 5 5 5 4 3 3 4 4 6 7 6 5 4 4 4 4 5 Steam total I, xlO Ib/ h 123,040 128,980 156,150 (continued) ------- TABLE 5-3. (continued) en i oo Time 2:30 2:45 3:00 3:15 3:30 3:30 3:45 4:00 4:07 4:15 4:30 4:45 5:00 5:15 5:30 5:30 5:35 5:45 6:00 6:15 6:20 6:20 Unit No. 1 Lower Upper chamber chamber temp., temp., •F 'C 1065 1155 1023 1148 1029 1148 1056 1147 Particle sizing 1107 1144 1120 1151 1115 1 1 50 Particle sizing 1102 1151 1091 1148 1072 1146 1057 1142 1056 1143 1038 1146 Particle sizing M5 outlet ended 1041 1148 1047 1150 1031 1150 Unit No. 2 Lower Upper chamber chamber temp., temp., •F 'C 961 1143 993 1137 1000 1138 1034 1150 outlet started 1045 1151 1050 1149 1095 1142 inlet started 1148 1134 1102 1138 1052 1140 1018 1144 1019 1143 1026 1150 outlet ended 1012 1146 996 1142 1010 1138 Power outage; inlet H5 and particle NOTE: at 6:20 p.m. power off due to Unit No. 3 Lower Upper chamber chamber temp., temp., •F 'C 1012 1028 1029 1072 1135 1162 1121 1074 1061 1149 1035 1037 1014 1025 1034 1008 sizing storm, 1148 1150 1143 1150 1143 1145 1142 1140 1144 1143 1144 1145 1141 1143 1145 1148 ended . ft Unit No. 4 Lower chamber temp., •F 1046 1026 1047 1073 1134 1164 1121 1085 1097 1099 1103 1104 1088 1102 1121 1080 Upper chamber temp., •c 1149 1136 1137 1148 1148 1143 1142 1142 1144 1143 1141 1140 1146 1145 1146 1145 Steam temp., •F 430 435 430 430 430 430 430 430 430 430 430 430 430 430 430 430 Steam Steam pressure, flow, psl Ib/h 355 360 355 355 355 355 355 355 350 350 350 350 350 350 350 350 63,000 65,000 64,000 64,000 64,000 63,000 64,000 64,000 61,000 61,000 61,000 61,000 62,000 61,000 62,000 62,000 1.0. fan, amps 200 200 200 200 200 200 200 200 200 200 200 200 200 200 200 200 Steam total Opacity, xlO Ib/ % h 4 3 3 4 4 4 5 5 5 4 3 3 3 3 3 3 190,656 ------- TABLE 5-4. en i PROCESS DATA FOR RUN 2— TUSCALOOSA ENERGY RECOVERY 22, 1985 Time a.m. 8:30 8:45 9:00 9:10 9:15 9:30 9:45 10:00 10:15 10:30 10:45 11:00 11:15 11:30 Unit No. 1 Lower Upper chamber chamber temp., temp., •F "C 1110 1150 1095 1148 1073 1148 Unit No. 2 Lower Upper chamber chamber temp., temp., •F 'C 1074 1140 1062 1152 1046 1159 Unit No, . 3 Unit No, . 4 Lower Upper Lower Upper chamber chamber chamber chamber Steam temp., temp., temp., temp., temp., •F 'C *F *C *F 1030 1028 1024 1144 1145 1146 1072 1092 1114 1155 1151 1148 355 355 355 Steam Steam pressure, flow, psi Ib/h 435 435 430 65,500 64,000 62,500 I.D. fan, amps 200 200 200 Steam total Opacity, xlO Ib/ % h 1 279,160 2 2 M5 inlet and outlet started 1063 1154 1061 1150 1071 1148 1108 1150 1097 1151 1077 1148 1064 1147 1033 1152 1046 1148 1032 1154 1102 1141 1086 1153 1052 1131 1035 1141 1054 1142 999 1142 1014 1143 1084 1140 1115 1150 1023 1142 Port change at 11:40 a.m.; cracked 11:45 p.m. 12:08 12:30 12:38 12:45 12:54 1:00 1:15 1:30 1:45 2:00 2:15 2:30 2:38 1076 1150 Particle sizing 1100 1148 Particle sizing 1029 1159 H5 inlet started 977 1145 977 1148 962 1150 962 1 161 955 1 1 50 974 1140 992 1 1 50 Particle sizing 1086 1148 outlet started 980 1139 inlet started 960 1139 932 1144 913 1141 892 1138 903 1142 929 1141 927 1145 932 1142 inlet ended 1042 1046 1114 1148 1148 1066 1048 1030 1043 1082 probe noted 1052 1017 1020 1013 1002 991 986 969 961 971 1124 1147 1146 1143 1144 1144 1148 1144 1143 1156 on inlet; 1144 1148 1151 1143 1143 1142 1149 1142 1143 1147 1084 1100 1080 1070 1072 1089 1090 1108 1128 1102 samp 1 i ng 1101 1040 1027 1058 1051 1047 1064 1059 1078 1111 1142 1156 1145 1159 1148 1148 1146 1145 1155 1150 stopped 1150 1151 1144 1147 1163 1179 1152 1146 1146 1146 350 350 350 350 350 350 350 350 350 350 355 360 360 360 360 360 360 360 355 355 430 430 430 430 430 430 430 430 430 430 430 435 435 435 435 435 435 435 435 435 58,000 60,000 61,000 62,000 58,000 60,000 58,000 58,000 57,000 59,000 64,000 66,000 64,500 65,000 66,500 67,500 67,500 66,500 65,500 62,500 190 195 195 195 205 200 200 195 200 200 200 205 210 215 215 215 215 215 215 220 2 3 3 3 3 3 2 3 3 4 4 1 1 2 2 1 1 1 2 2 (continued) ------- en O TABLE 5-4. (continued) Time 2:45 3:00 3:08 3:15 3:20 3:30 3:45 4:00 4:02 Unit No. 1 Lower Upper chamber chamber temp., temp., •F "C 1023 1150 1069 1151 Particle sizing 1047 1150 M5 outlet ended 1043 1151 1025 1148 1061 1152 M5 inlet ended Unit No. 2 Lower Upper chamber chamber temp., temp.. *F 958 987 outlet ended 1017 1018 1029 1062 •c 1146 1150 1143 1150 1142 1134 Unit No. 3 Lower chamber temp., •F 1010 1035 1090 1065 1020 1039 Upper chamber temp., •c 1145 1144 1146 1145 1141 1143 Unit No. 4 Lower chamber temp.. •F 1185 1256 1204 1037 1114 1124 Upper chamber temp., •c 1145 1140 1147 1146 1149 1144 Steam temp.. •F 355 355 355 350 350 350 Steam Steam pressure, flow, psi 435 435 435 430 430 430 Ib/h 63,500 63,500 61,000 58,000 58,000 58,000 I.D. Steam total fan, Opacity, xlO Ib/ amps 210 205 200 200 200 195 % 3 3 3 3 4 3 h 321 ,790 ------- TABLE 5-5. PROCESS DATA FOR RUN 3—TUSCALOOSA ENERGY RECOVERY May 23, 1985 Time a.m. 8:00 8:30 8:40 8:41 8:45 9:00 9:30 10:00 10:15 10:30 10:40 10:45 :00 :15 :30 :40 :45 :58 p.m. 12:00 12:15 12:30 12:45 :00 :15 :30 :40 :45 2:00 2:15 2:27 Unit No. 1 Unit No. 2 Lower Upper Lower Upper chamber chamber chamber chamber temp., temp., temp., temp., •F *C *F "C 1109 1146 1001 1136 1078 1149 1037 1126 M5 outlet started M5 inlet started 1092 1147 1050 1143 1105 1146 1063 1133 1119 1150 1034 1143 1066 1150 1064 1139 1052 1141 1054 1140 1036 1150 1038 1138 Particle sizing outlet started 1003 1151 1082 1137 1001 1150 1131 1139 993 1150 1163 1142 1043 1148 1117 1150 M5 outlet ended 1068 1147 1110 1142 Particle sizing Inlet started 1085 1153 1103 1139 1173 1147 1057 1142 1192 1146 1088 1143 1214 1146 1124 1142 1169 1148 1085 1143 1094 1150 1033 1140 1078 1148 1018 1143 Particle sizing outlet ended 1054 1152 971 1139 1037 1149 1020 1125 1006 1149 1011 1149 M5 inlet ended Unit Lower chamber temp., •F 1087 1031 1019 1009 1004 996 996 1003 987 982 975 986 1003 1007 1152 1122 1084 1043 998 995 983 1015 973 No. 3 Upper chamber temp., •c 1144 1144 1144 1144 1143 1148 1146 1150 1145 1146 1145 1143 1147 1147 1145 1144 1142 1144 1146 1141 1147 1135 1148 Unit Lower chamber temp., •F 1014 1051 1126 1090 1126 1092 1080 1063 1034 1025 1015 1023 1022 991 1094 1081 1063 1018 1006 1036 1015 1012 1052 No. 4 Upper chamber temp., •c 1152 1157 1155 1145 1150 1142 1148 1142 1150 1152 1155 1142 1144 1146 1144 1142 1142 1152 1146 1148 1146 1148 1146 Steam temp., •F 435 435 435 435 435 435 435 435 435 435 435 435 435 435 435 435 435 430 430 430 430 430 430 Steam pressure, PSl 355 355 355 350 350 355 355 360 360 360 360 360 360 360 350 350 350 350 350 350 350 350 350 Steam flow, Ib/h 65,000 64,500 64,500 63,000 62,000 65,000 65,000 66,50p 66,000 66,000 66,500 65,000 65,000 65,500 62,000 62,000 62,000 58.000 54,500 56,000 57,000 58,500 60,500 I.D. fan, amps 215 215 215 215 215 215 215 215 215 215 215 215 215 215 215 215 215 205 200 200 195 195 195 Opacity, 1 2 2 2 2 3 4 3 7 5 2 2 2 3 3 3 7 Steam total xlO Ib/ h 419,700 457,450 (continued) ------- TABLE 5-5. (continued) C71 ro Unit No. 1 Unit No. 2 Time 2:30 2:45 3:00 3:12 3:15 3:30 3:45 4:00 4:15 4:30 4:42 5:00 Lower Upper Lower chamber chamber chamber temp., temp., temp.. *F 'C "F 1007 1148 1023 1009 1147 1037 1025 1148 1005 Upper chamber temp.. •c 1146 1142 1142 Unit No. 3 Lower chamber temp., •F 989 1008 996 Upper chamber temp., •c 1147 1148 1149 Unit No. 4 Lower chamber temp.. •F 1050 1048 1079 Upper chamber temp.. •c 1146 1145 1148 Steam temp., •F 430 430 430 Steam pressure psi 350 350 350 Steam , flow. Ib/h 58,500 57,000 56,500 I.D. Steam total fan. Opacity, xlO Ib/ amps 195 195 205 % 5 4 5 h Particle sizing inlet started 1035 1151 1013 1034 1147 1054 1019 1146 1034 996 1146 1018 1007 1150 1046 1029 1151 1030 1137 1139 1144 1147 1145 1144 1025 1101 1082 1066 1057 1067 1142 1143 1145 1147 1142 1145 1064 1098 1077 1059 1012 1020 1147 1147 1148 1149 1145 1131 430 430 430 430 430 430 350 350 350 350 350 350 55,000 58,000 58,000 58,000 58,000 58,500 200 195 195 195 195 195 6 5 3 3 3 4 Particle sizing inlet stopped 474,873 ------- The average ESP power level during Run 2 was estimated to be 2.0 kVA, repre- senting 7 percent of the design power level. During Run 3, the outlet field tripped off from 11:00 to 11:30 a.m., and was restarted at 11:30 a.m. The average ESP power level during Run 3 was estimated to be 5.7 kVA, repre- senting 21 percent of the design level. Tables J.3-1, J.3-2, and J.3-3 in Appendix J present the recorded values of the monitored ESP parameters during Runs 1, 2, and 3, respectively. 5.4 SUMMARY OF PROCESS OPERATIONS DURING TESTING The incinerator modules were operated under steady and normal conditions at approximately 90 percent of capacity throughout the tests. It is expected that the inlet sampling data are representative for uncontrolled emission levels. (Approximately 90 percent of all modular plants are uncontrolled.) The ESP operation was not steady and the ESP power levels represent only a small fraction of the design power level. (This ESP as well as the control devices on the modular plants with controls has a design removal efficiency of only approximately 50 percent.) Air inleakage at the ID fan flange occurred throughout most of the test period. It is expected that the outlet sampling results do not reflect representative conditions for controlled emission levels. The inlet chromium emission levels will be used to represent the 90 percent of the modular facilities without controls. By applying a removal efficiency of 50 percent to the inlet chromium emission levels, typical chromium emissions for the uncontrolled facilities can be estimated; thus, the suspect outlet data from the Tuscaloosa facility will not have to be used to represent current controlled emission levels. 5-13 ------- The temperature .in the primary chamber of each incinerator is maintained between 540° and 760°C (1000° and 1400°F). Secondary chamber temperatures are typically 1150°C (2100°F). No auxiliary fuel is used during normal operation. Approximately 99 percent of the refuse incinerated at the Tuscaloosa facility is supplied by residential sources in eight nearby municipalities. The remaining 1 percent [4.5 Mg/day (5 tons/day)] consists of scrap tires from the nearby tire plant. Large metal items are sorted from the refuse prior to incineration and sent to a landfill. Exhaust from the incinerators is routed through a common ESP for par- ticulate emission control prior to exiting through a single stack. The ESP was manufactured by the PPC Company of Longview, Texas. It has two fields with 14 plates in each field. The plates are spaced approximately 25 cm (10 in.) apart and are 7.6 m (25 ft) long and 2.4 m (8 ft) wide. The applied voltages to the first and second fields are 31 and 30 kV, respectively. The specific collection area is estimated to be 400 m2/1000 acmm (120 ft2/1000 acfm). The ESP has a design removal efficiency of 50 percent and collects approximately 320 kg (700 Ib) of fly ash per day. All tests were conducted while the incinerators were operating under normal conditions. However, the ESP experienced some operational difficul- ties at various times throughout the test program. 5-14 ------- |