SWRHL-IO!
      THE DESIGN AND CONSTRUCTION OF A LASER METER
FOR THE MEASUREMENT OF CW IRRADIANCE AT 623.8 NANOMETERS
                           by
                Richard W. Peterson, M.S.
                Wi Ibur  F. Van Pelt, M.S.
                Harold  F. Stewart, Ph.D.
               Electronic Products Program
       Southwestern Radiological Health Laboratory

   U.S. Department of Health, Education, and Welfare
                 Public Health Service
              Environmental Health Service
                      November 1970

-------
                                                          SWRHL-IOI
      THE DESIGN AND CONSTRUCTION OF A LASER METER
FOR THE MEASUREMENT OF CW IRRADIANCE AT 623.8 NANOMETERS
                           by
                Richard W. Peterson, M.S.
                WiIbur  F. Van Pelt, M.S.
                Harold  F. Stewart, Ph.D.
               Electronic Products Program
       Southwestern Radiological Health Laboratory

    U.S. Department of Health, Education, and Welfare
                  Public Health Service
               Environmental  Health Service
           Environmental Control Administration
               Bureau of Radiological Health
                     November 1970

-------
                        ABSTRACT

A description of the design and construction of a meter capable
of measuring scattered levels of irradiance (power density) re-
sulting from the operation of low power, continuous wave, helium-
neon gas lasers is presented.  Readily available components are
used and the techniques for calibration and use of the instrument
are discussed.  The instrument responds to irradiances of 10 nW/c
or greater depending on the aperture size selected by the user.

-------
                    ACKNOWLEDGEMENTS
The authors would like to express their appreciation for
the assistance of Mr.  Mason MeNinch in the testing and
evaluation of photo conductive detectors and for his aid
in constructing the prototype.
The efforts of Miss Anne Roberts in the preparation of
figures and in reviewing the text are gratefully acknowl-
edged.
                          ^^

-------
                         TABLE OF CONTENTS

                                                                  Page
ABSTRACT                                                            i
ACKNOWLEDGEMENTS                                                    i i
LIST OF FIGURES AND TABLES                                          iv
INTRODUCTI ON                                                        I
TECHNICAL CONSIDERATIONS                                            I
CONSTRUCT I ON                                                        3
    I.  Detector Circuit                                            3
    2.  Optical System                                              5
    3.  Assembly                                                    7
CALIBRATION AND OPERATION                                           7
DISCUSSION                                                         I I
CONCLUSIONS                                                        12
REFERENCES                                                         14
APPENDIX A                                                         15

-------
                         LIST OF FIGURES






Figure                                                             Page




I.  Basic Detector Circuit                                           3




2.  Complete Detector Circuit                                        4




3.  Optical System                                                   5




4.  Meter Response vs. Angle of Incidence                            6




5.  External View of the Assembled Meter                             8




6.  Internal View of the Meter                                       8




7.  Arrangement of Apparatus for Calibration                         9




8.  Response of the Meter vs. Relative Irradiance                   10




9.  Spectral Response                                               13








                          LIST OF TABLES






Table                                                              Page




I.  Aperture Diameters and Areas                                     6
                                 ^V

-------
                           INTRODUCTI ON

The  increasing availability and  use of  low power helium-neon  laser
units has created  situations which demand the measurement of  irradi-
ance  (power per unit area or power density)  in order to adequately
assess any possible ocular  hazards.  Commercially available instru-
mentation is typically  designed  to provide a  large dynamic range of
power measurement.  Since ocular hazard analysis requires low  level
irradiance determination, we have designed, constructed and tested an
instrument which determines irradiance directly at the helium-neon
laser wavelength,  632.8 nm  (nanometers), over a narrow dynamic range
of  irradiance values centered about that value considered to denote
the maximum allowable irradiance..

                    TECHNICAL CONSIDERATIONS

Lasers produce a highly collimated, coherent beam of light.    In the
visible region of  the electromagnetic spectrum the human eye  is the
organ of critical  concern.  Relatively  low irradiance levels can be
concentrated by the focusing action of the cornea and the lens of the
eye, which result  in irradiances on the retina on the order of 100,000
times greater than the cornea I  irradiance.   In evaluating ocular hazards
one  is interested  only  in the measurement of potentially hazardous
exposure levels to the exterior  of the eye, and detection instrumentation
must be designed to respond reliably to the cornea I  (or low level) irradi-
ance; response to  those high levels of  irradiance which will  produce
observable lesions on retinal  tissue is then not necessary.   Since direct
laser beam levels  generally exceed (by several orders of magnitude)
(acceptable)  corneal  levels, measurements will routinely be made on
scattered levels.  An inexpensive photoconductive detector was selected
for use in the instrument because of its high sensitivity, its speed of
                                I

-------
response as well as the success achieved in its use in commercially
produced photographic tight meters.  While the detector could have
been used to provide a measurement of power only, the measurement of
irradiance was achieved through the use of an aperture of known
dimensions. To minimize the effects of any uneven responses across
the surfaces of the detector, a lens and diffusing screen were neces-
sary to provide uniform surface illumination.   Finally, the response
of the detector to a  large range of wavelengths required a filter
between the aperture and detector which passes only those wavelengths
at or near the wavelength of interest (632.8 nm).

At this time the U.S. Public Health Service does not advocate or endorse
any published set of  laser exposure standards, guides or recommendations.
The design of  the prototype meter was therefore based on the premise
that an internally adjustable control would be necessary to center the
response of the instrument at any  irradiance which might be typical of
the various published recommendations.  This control can be eliminated
at such time that a single hazard  irradiance  level  is adopted.

Experience with optical power meters has indicated that alignment of a
detector for maximum readings is very difficult if the only indication
is on a meter, since the operator must devote his attention to two
points simultaneously.  Since the meter is intended for survey applica-
tions, its design includes an audible indicator to allow the surveyor
to devote his attention solely to the alignment of the detector with
respect to the radiation to be measured.

To achieve maximum utility, the meter, which is capable of field oper-
ation, has a minimum weight and no external  power requirements.   The
size and ruggedness is such that the entire instrument can be operated
and held in one hand.

-------
                           CONSTRUCTION

   I.  Detector Ci rcu it:
      The detector circuit,  in  its most elementary form,  is the series
  circuit shown  in Figure  I.  The conductance of the detector is pro-
  portional to the total power of the light incident on the photo-
  conductive surface, which  is deposited on an inner surface of the
  detector.  Thus, the current  in the meter is related to the incident
   Iight power.
                               photoconductive
                                     detector
        battery
                                       meter
Figure 1.  BASIC DETECTOR CIRCUIT
     The  complete  detector  circuit  schematic  is shown  in Figure 2.  To
     provide  a  constant  voltage  source,  a  limiting  resistor, R|, and
     Zener diode have been  added.   The  Zener  diode  regulates the

-------
   potential  between points "a" and "b" at 10 volts as long as the
   battery voltage remains slightly above this level.  Failure of
   the battery,  B,,  can  be monitored by constructing a voltmeter
   circuit which can be  introduced across points  "a" and "b" by
   S-2, a push-button switch.   The variable series resistor, R2,
   can be set to provide a full-scale indication  when the batteries
   are new and the potential  "a" to "b" is 10 volts.  Any subsequent
   test of battery indication  which results in less than a full-scale
   deflection shows  battery degeneration and  indicates the need for
   replacement of the meter supply batteries.   The variable parallel
   resistor,  R3, provides a method of adjusting the meter response to
   the appropriate point such  that the midpoint of meter deflection
   will indicate that level of irradiance considered to be hazardous.
Figure 2.  COMPLETE DETECTOR CIRCUIT
   R-l
                 S-l
                         S-2
                                                       CL-505

-------
    Selection  of  a  meter with internal relay points  (adjustable  from 0
    to full-scale triggering) was based on the need  for  an  alternate
    indicator  mode.   This mode is an audible signal  furnished  by  a  small
    oscillator and  speaker which produces a signal at about 3,000 Hz.   A
    second  battery,  62,  is used to power this device.  The  operability
    of this segment of the circuit is easily verified by adjusting  the
    relay  trigger point to zero deflection on the meter.

    2.   Optical  System:
         As  noted  earlier, the hazard meter must be designed to respond
    to  irradiance (power density) and not to total optical  power  at the
    point  of interest.  This is achieved by using the optical  system
    shown  in Figure 3.  Although the aperture must be of known size,
    the  section of  the size  is dependent only on:
          I) the uniformity of irradiance across the  aperture and,
         2) the sensitivity  required of the system.
Figures.  OPTICAL SYSTEM
all dimensions in millimeters
 (machined parts aluminum)
                                         instrument case
                                                CL-505 photoconductive detector
                                          diffusing glass

                                          filter

-------
   Several apertures  have  been used and their  characteristics are
   Ii sted  in Table  I:

                                TABLE  I
                     APERTURE DIAMETERS AND  AREAS
Aperture No.

1
2
3

4
Di

1 .
1 .
0.

0.
ameter

46 cm
1 2 cm
56 cm

21 cm
Area

1 .
0.
0.

0.
2
67 cm
2
99 cm
25 cm2
2
035 cm
   All  internal  surfaces of the collimator  have been sprayed with  a
   velvet  black diffuse paint to minimize  internal  reflections.  The
    lens  is  positioned so that the focal  point  falls at the diffusing
   glass.   This provides a well defined  acceptance angle and eases
   the  problem of spurious responses from  light entering the detector
   from  directions not along the axis of the  light beam to be measured.
   The  angular acceptance width to the half maximum response points
    (FWHM as indicated  in Figure 4)  is approximately  12  for all
   apertures evaluated.
APERTURE NO. i
                APERTURE NO. 2
                                   APERTURE NO. 3
                                                    APERTURE NO. 4
 0    '"   20 «   10   o   10  20
ANGLE              ANCLE
                                                jo  20   10
                                                          ANGLE
              Figure 4. METER RESPONSE vs. ANGLE OF INCIDENCE

-------
3.  Assembly:
    The assembled configuration of the  laser hazard meter is shown
in Figures 5 and 6.  The system is contained in a 5-inch  long by
4-inch wide by 3-inch deep aluminum box.  No attempt was made in
this prototype to utilize miniature components and it is apparent
that even this small size could be further reduced.  The  internal
controls, R2 and R-j, are accessible through holes on the bottom of
the case.  In operation, these holes should be covered to reduce
stray  light  leakage to the detector.

                  CALIBRATION AND OPERATION

The assembled laser hazard meter is calibrated by utilizing an ex-
panded beam from a 2-mW helium-neon laser.  Since the irradiance of
the cross sectional plane of the beam is approximated by a gaussian
distribution, expansion of the beam is necessary to assure a nearly
constant  irradiance across the aperture.  A calibrated power meter,
fitted with a variable iris, is used to determine the irradiance
at the aperture plane.  The beam is then reduced to the irradiance
                                              2
assumed to be hazardous (for example, 10 yW/cm ) by inserting cali-
brated optical filters in the beam.  A reference power meter,
coupled to the system by a beam splitter, monitors the stability of
the laser output during the calibration procedure.  The arrangement
of the calibration apparatus is shown in Figure 7.

The resistor, R3, is then adjusted to give half-scale response.   The
linearity of response can be checked by varying the filters in the
beam and noting the response of the laser meter (Figure 8), although
this is not of great importance since the meter is a single point,
threshold device.

-------
Figure 5. EXTERNAL VIEW OF
        THE ASSEMBLED METER
  Figure 6.  INTERNAL VIEW OF
          THE METER
               8

-------
Figure 7.   ARRANGEMENT OF

            APPARATUS FOR

            CALIBRATION
                                         meter to be calibrated
                                                         calibration plane
          beam steering


           mirrors
                                reference power meter
         \
   A.
beam splitter

-------
90
80
70
60
50
40
30
20
10
       Figure 8. LASER METER READING vs.
               RELATIVE IRRADIANCE
   P
   5
 0
                  .3      .4     .5
                LASER METER READING
.6    .7
                                              .8
.9
                       10

-------
Finally, the battery metering adjustment, R-,,  is adjusted for
full-scale deflection.  This calibration procedure  requires the
use of a reasonably well-equipped  laboratory.

In  practice, the operator aligns the meter with the beam of light
at the position of  interest.  By setting the relay  contacts to
trigger at half-scale, the sonic alert will sound at any time the
irradiance exceeds the hazardous  level.

The operational  life of the meter circuit batteries should be a
minimum of  10  hours as based on the battery manufacturers' speci-
fications  if the unit  is operated at normal room temperature.
Since measurements are generally of short duration, several months
of routine operation will be obtainable before the  batteries must
be replaced.   Replacement of the cells in the  field (facilitated
by the use of  readily available 9V transistor  radio batteries)
will not change the instrument calibration because of the Zener
diode regulator circuit.

                       DISCUSSION

Use of the laser meter has proven to be straightforward.  The prime
restriction that the entire aperture be uniformly illuminated
remains.   This restriction becomes of lesser importance as the
aperture area decreases.   Only beams having diameters less than
the aperture cannot be measured with the meter.  Since this condi-
tion generally exists only in the direct beam  (which is assumed to
be dangerous) measurement can be made of most  scattered light (from
either diffuse or specular surfaces).  Specular reflections from
flat surfaces retain the divergence characteristics of the incident
beam and must be analyzed by other techniques.

-------
Aperture dimensions used In the prototype were randomly selected.
If one were to utilize the meter for measurement of the ACGIH
                                                 2
guide levels, for example, of 50, 20 and 10 yW/cm  (based on
ambient lighting conditions) then three apertures of relative
areas 0.2, 0.5 and 1.0 might be used.   It is then necessary to
                                         2
calibrate at only one level (say 10 yW/cm ) and by appropriate
choice of aperture any of the other guide levels would also be
caIi brated.

                                                                  (2)
The detector selected has a spectral response as shown in Figure 9
When the unit is calibrated at 632.8 nm it should not be used for
other wavelengths due to the rapid change in detector response as
a  function of wavelength.  The detector has held its calibration
with no measurable change over a period of 60 days.  Tests  in which
the detector was continuously  illuminated over an eight-hour period
           2
by 20 yW/cm  at 632.8 nm failed to produce any calibration shift.
The detector exhibits a change in conductance of 7 percent over a
temperature  range of 0°C to 50°C.

Ambient  light does contribute  to the meter response.  Typical room
illumination contributes between 5 to 10 percent of scale response,
which has  been found to be an  insignificant contribution because of
the nonlinearity at  low  levels.
                             CONCLUSIONS

The feasibility of a portable instrument suitable for detecting
hazardous  levels of continuous wave laser radiation at 632.8 nm
has been demonstrated.  Utilization of the instrument provides a
convenient method of detecting hazardous levels of reflected or
scattered  Iight.

The instrument  is stable, easy to use and can be constructed from
readily available electronic components.

-------
100
 90
 80
 70
 60
 50
 40
 30
 20
 10
               Figure 9.
                 SPECTRAL RESPONSE
                      (reference 2)
   400
500           600           700

       WAVELENGTH (nanometers )

                13
800

-------
The instrument exhibits those properties of cost, utility, and
simplicity of operation which indicate its usefulness for
relatively unskilled field survey personnel.

References:

(I)  A Guide  for Uniform Industrial  Hygiene  Codes or Regulations
     for Laser Installations, Issued  by The American Conference
     of Governmental Industrial  Hygienists, 1968.

(2)  Photoconductive Cell  Designers Kit,  Essential  Technical  Data,
     Clairex Corporation,  8 W.  30th St.,  New  York,  New York  10001,

-------
                       APPENDIX A


COMPONENTS* USED IN THE LASER METER

     Detector:  Photoconductive cell, CI airex CL-505
                  AlI ied Radio 60 D 7458  (I.50)

     Batteries (3):   NEDA I604D (9V).  Allied Radio 18 D
                          5769 (2.05)

     Battery Clips  (3):   Allied Radio 18 D  5184 (4.20)

     Resistors:   R-I    500 ohm, 1/2 watt, 20  percent fixed (0.36)
                 R-2   10,000 ohm,  2 watt,  variable (1.60)
                 R-3   1,000 ohm,  2 watt, variable (1.60)

     Switches:  S-I   SPST, Push Button,  Black,  Allied  Radio
                        56 D 9947  (0.65)

                S-2   SPOT, Push Button,  Red,  Allied Radio
                        56 D 4946  (0.65)

     Diode:   D-I,  10 V  Zener reference Diode, Type  N758 (1.00)

     Meter:   M-I, 0-1 mA DC, internal magnetic  relay,  API
                Instruments Model  202M.   A  lied Radio  51 D
                5584 (37.00)

     Sonic Alert:   Mai lory Sonalert Model SC628.  Allied Radio
                     60 D 8983 (5.50)

     Case:   Aluminum Minibox.   Bud  No. CU02I05-A.   Allied Radio
              42 D  7621  (I.50)

     Lens:   Piano Convex.   Diameter 19.5  mm,  Focal  length 38 mm
              Edmund  Scientific No.  94,035  (0.60)

     Optical  Housing:   Machine from stock (see  Figure  3).
              (cost  variable)

     Filter:   I" x  I",  cut from Rubylith  filter gel.

     Diffusing glass:   I" x I", but from 4" x 6" piece of 1/8"
                       opal  glass.   (1.00)
     Paint:   3-M NexteI  Velvet Black (2.40  per  can)

     Equivalent  components may be  substituted for any  of the models
     specifically designated above.
* Mention of commercial  products used in connection with work reported
in this article does not constitute an endorsement by the Public Health
Servi ce.
                             15

-------
                                  DISTRIBUTION

 1  - 20  SWRHL, Las Vegas, Nevada
     21  John C.  Vi11 forth,  Director,  BRH,  Rockville,  Maryland
     22  E.  C.  Anderson,  BRH.  Rockville,  Maryland
     23  Director,  Criteria  &  Standards,  BRH.  Rockville,  Maryland
     24  Director,  Grants  Office,  BRH,  Rockville, Maryland
     25  Robert H.  Neill.  Program  Office, BRH, Rockville. Maryland
     26  John G.  Bailey, Office of Information, BRH, Rockville, Maryland
     27  Roy  H. Rohn, Jr., Administration Management,  BRH, Rockville. Maryland
     28  James  W. Miller.  Regional Operations, BRH, Rockville, Maryland
     29  William  A. Mills, Div. of Biological Effects, BRH, Rockville, Maryland
     30  George E.  Anderson. Chief, Radiation Physics  Lab, BRH, Rockville. Md.
     31  Harry  D. Youmans. Div. of Biological Effects. BRH. Rockville, Maryland
     32  Arve H.  Dahl, Medical Radiation Exposure. BRH. Rockville, Maryland
33  - 34  Charles  L. Neaver, Div. of Environmental  Rad., BRH, Rockville, Md.
     35  Robert Elder, Div. of Electronic Products, BRH, Rockville, Md.
     36  Roger  Schneider,  Rad. Measurement & Cal.  Br., BRH, Rockville, Md.
     37  Lavert Seaborn, Intelligence Br., BRH, Rockville, Md.
     38  Walt Gundaker, Product Testina & Evaluation Br. , BRH, Rockville, Md.
     39  Bob  Britton, Compliance Br., BRH, Rockville,  Md.
     40  William  Properzio, X-ray  Exposure Control Lab., BRH, Rockville, Md.
     41  George R.  Schultz, Training Inst., ECA, Rockville, Md.
     42  Elizabeth  Boeker, Assoc.  Dir., ECA, Rockville, Md.
     43  C. Bruce Lee, Office Research & Develop., ECA, RocKvilie, Ma.
     44  FredericK  Erickson, Training & Manpower Develop., ECA, Rockville, Md.
     45  Mark H.  Barnett,  Rad. Health Br., ECA, Rockville, Md.
     46  Charles  Froom, Rockville  Training Sec., ECA,  Kockville, Md.
     47  Richard  Marland,  Program  Development, ECA, Rockville, Md.
     48  Robert M.  Hallisey, Regional Rep., Region  1,  Boston, Mass.
     49  Saul J.  Harris, Region  II Regional Rep., New  York, N. Y.
     50  Robert Frankel, Region  III  Regional  Rep.,  Philadelphia,  Pa.
     51  Joseph Brennan,    Region  IV  Regional  Rep.,  Atlanta,  Ga.
     52  Jerome A.  Halperin,  Region  V Regional  Rep.,  Chicago,  111.

-------
Distribution - continued

53  James A. McTaggart,  Region VI Regional  Rep.,  Dallas, Texas
54  Gerald A. Jacobson,  Region VII Regional  Rep., Kansas City, Mo.
55  Paul B. Smith, Region VIII Regional  Rep., Denver, Colorado
56  Clifford E. Nelson,  Region IX Regional  Rep.,  San Francisco, Calif.
57  Edward J. Cowan, Region X Regional  Rep., Seattle, Wash.
58  Morgan S. Seal, NERHL, Winchester,  Mass.
59  George Coates, Electronic Products,  NERHL, Winchester, Mass.
60  Neil Gaeta, Training, NERHL, Winchester, Mass.
61  Library, NERHL, Winchester, Mass.
62  C. R. Porter, SERHL, Montgomery, Alabama
63  Charles Phillips, SERHL, Montgomery, Alabama
64  Bob Callis, Office of Education & Information, SERHL, Montgomery, Ala.
65  Ralph E. Shuping, Physical Science Unit, SERHL, Montgomery, Ala.
66  Library, SERHL, Montgomery, Ala.
67  Richard Tells, N Y University Medical Center, Tuxedo, N Y
68  Wendell McCurry, Public Health Eng.II,  Bur. Environ.Health, Reno, Nev.
69  Mark Jakobson, Physics Dept., Univ.  of Montana, Missoula, Mont.
70  Albert Shephard, Georgia Inst. of Technology, Atlanta, Ga.
71  Oliver D. T. Lynch,  Jr., Rad. Safe.  Spec., Rad.Safe.Br., AEC, Mercury, Nev.
72  Paul Bolton, REECo Indust.Safety, Mercury, Nevada
73  Leroy Garcia, REECo. Indust. Safety, Mercury, Nevada
74  D. W. Hendricks, Radiological Operations, AEC/NVOO, Las Vegas, Nevada
75  Henry J. L. Rechen,  Div. of Electronic Products, BRH, Rockville, Md.
76  William Link, BRH Library, Rockville, Maryland
77  D. Hamil, AEC/NVOO,  Library, Las Vegas, Nevada

-------