EPA-R2-72-099
                                                68-02-0309/4-72
 IN-STACK TRANSMISSOMETER TECHNIQUES
      FOR MEASURING OPACITIES OF
         PARTICULATE EMISSIONS
       FROM STATIONARY SOURCES
                    by
           Carl M. Peterson, Ph.D.
             M. Tomaides, Ph.D.
ENVIRONMENTAL RESEARCH CORPORATION
          3725 North Dunlap Street
          St. Paul, Minnesota 55112
                Prepared for

  ENVIRONMENTAL PROTECTION AGENCY
    National Environmental Research Center
       Division of Chemistry and Physics
           Research Triangle Park,
           North Carolina 27711
           Contract No. 68-02-0309
                 April 1972

-------
  IN-STACK TRANSMISSOMETER TECHNIQUES
      FOR MEASURING  OPACITIES  OF
         PARTICULATE  EMISSIONS
        FROM STATIONARY  SOURCES
                   by
        Carl M,  Peterson,  Ph.D.
           M. Tomaides,  Ph.D.
  ENVIRONMENTAL  RESEARCH CORPORATION
       3725 North Dunlap Street
      St. Paul,  Minnesota   55112
             Prepared  for

    ENVIRONMENTAL  PROTECTION AGENCY
National Environmental  Research Center
   Division of  Chemistry and Physics
        Research Triangle Park,
         North  Carolina 27711
        Contract  No.  68-02-0309
               April  1972

-------
                       ABSTRACT
Field studies were  conducted to obtain basic research  data
as a base for developing design and performance  specifica-
tions for transmissometers which are to be used  to  measure
smoke stack plume opacities.  Tests, conducted on the  stack
of a pulverized  coal-fired power plant, were designed  to
evaluate the influence  of transmissometer illumination and
light receiving  angles,  and transmitted light wavelength on
in-stack opacity measurements and their correlation with
the stack plume  opacity.

Two specially-designed  transmissometers, one having a
small fixed illumination-viewing angle design and the  other
having adjustable illumination and viewing angles,  were
mounted on a cylindrical 145-inch diameter steel stack to
measure the in-stack  opacity,  A 0.5° telephotometer was
used to determine the out-of-stack opacity of the plume as
viewed from a distant river bank.

The results show a  significant dependence of measured  in-
stack transmittance as  a function of illumination and
receiver angles.  This  dependence is most pronounced at
small angles.  The  measured transmittance increases with
increasing illumination  and viewing angle, resulting in
5 and 46 percent errors  for 53 and 60° angles, respectively,
when compared to true transmittance.  The dependence of
measured transmittance  as a function of illumination light
wavelength was also established.

Using a small fixed illumination^viewing angle design
transmissometer,  a  good  correlation was obtained between
plume opacity and in-stack transmittance.

This report was  submitted in fulfillment of  Contract
68-02-0309 under the  sponsorship of the Environmental
Protection Agency,
                          111

-------
                        CONTENTS







Section                                               Page







 I           CONCLUSIONS                                 1





 II          RECOMMENDATIONS                             3





 III         INTRODUCTION                                5




 IV          OBJECTIVES                                  9





 V           INSTRUMENTATION                            11





 VI          LABORATORY EVALUATION  OF  INSTRUMENTS      21





 VII         FIELD FACILITIES AND CONDITIONS           33





 VIII        RESULTS  AND DISCUSSION                     45





 IX          ACKNOWLEDGEMENTS                           53





 X           REFERENCES                                 55





 XI          APPENDICES                                 57

-------
                        FIGURES

Number                                                   Page

   1      Schematic  and Basic Dimensions  of  the          12
          Reference  Transmissometer Transmitter

   2      Schematic  and Basic Dimensions  of  the          13
          Reference  Transmissometer Receiver

   3      Photograph of the Reference                    15
          Transmitter Assembly

   4      Schematic  of the Experimental Trans-           17
          missometer Transmitter

   5      Schematic  of the Experimental Trans-           19
          missometer Receiver

   6      Linearity  of the Reference Transmisso-        24
          meter  Readout as Determined Using
          Neutral  Density Light Filters

   7      Linearity  of the Experimental Trans-           30
          missometer Readout as Determined Using
          Neutral  Density Light Filters

   8      Experimental Stack Dimensions and              34
          Port Locations

   9      Experimental Steel Stack Viewed Against       36
          the Concrete Stack Used as the  Contrasting
          Target for the Telephotometer Measure-
          ment of  Plume Transmittance

  10      Example  of the Experimental Data               39
          Recorded in the Field Test

  11      Example  of the Experimental Data               40
          Recorded in the Field Test

  12      Transmittance Measurements by 3° Angle        46
          of View  and 1.5° Illumination Angle
          Transmissometer Inside Stack and 0.5°
          Angle  of View Telephotoraeter Outside
          Stack
                          VI

-------
                  FIGURES (continued)
Number
  13      In-Stack Experimental Transmittance        48
          Detected at Various Transmitter and
          Receiver Angles of the Experimental
          Transmissometer Normalized for  Constant
          True Reference In-Stack Transmittance
          of 0.807.

  14      Summary of Interference Filter  Tests.      51
          Relationship of the Experimental In-
          Stack Transmittance and Transmitter
          Angle for Four Interference Light
          Filters at 5° Receiver Angle.
                           Vi i

-------
                         TABLES
Number                                                Page

   1      Stability  of  Reference Transmissometer      25
          With  Varying  Line Voltage

   2      Stability  of  Reference Transmissometer      26
          During  Startup

   3      Dark  Current  Signal of Experimental         28
          Transmissometer Receiver

   4      Calculated Accuracy of Transmittance        32
          Measurements

   5      Data  on  the Plume Transmittance vs          45
          In-Stack Transmittance Correlation
                          VI 11

-------
                       SECTION  I
                     CONCLUSIONS


The major conclusions  drawn  from  the  test results are:

1.   The apparent transmittance of  an aerosol as measured by
a transmissometer increases  with  the  size of the transmitter
and receiver angles of  the transmissometer.   This is due to
the increase in the amount of  scattered  light detected by
the transmissometer as  these angles increase in size.
Detection of scattered  light by a transmissometer represents
an error in the measurement  and should be minimized as much
as possible .

2.   The measured transmittance can be singly influenced by
varying either the receiver  or transmitter angle and both
these angles are nearly equally influential  in effecting the
measured transmittance  within  0-60°.

3.   A positive error  of about 5  percent was noted when 5°
receiver and transmitter angles were  employed for transmit-
tance measurements of  a pulverized  coal-fired boiler effluent.
The error increased to  about 46 percent  when 60° receiver
and transmitter angles  were  employed.  The tests were con-
ducted with an experimental  transmissometer  using a tungsten
lamp, a photoelectric  detector, and a light  beam diameter
(aperture size) of approximately  four inches.

4.   The transmittance  of the  pulverized coal-fired boiler
effluent studied was light wavelength dependent.  A differ-
ence from 6 to 10 percent in transmittance was measured
when employing interference  filters of 0.438 to 0.656 micron.
The 0.656 micron filter (red band)  consistently provided the
lowest in-stack transmittance.  This  result  would indicate
that the particles in  the effluent  were  sufficiently large
to exceed the first maximum  of the  particle  extinction curve
and/or significant spectral  absorption effects were present.

5.   Good agreement was found  between the opacity of the plume
and the in-stack transmittance of the coal-fired boiler
emission.  For high transmittance conditions  (greater than
70 percent) ideal agreement  was observed between remote
plume transmittance measurements  by telephotometry of con-
trasting targets in back of  the plume and in-stack trans-
mittance measurements  by a transmissometer with a 3° receiv-
ing angle, 1.5° transmitter  angle,  and a tungsten lamp light

-------
diameter of 0,5 inch.  For lower  transmittance  conditions
(less than 40 percent!, the in-^stack  measurement  was
approximately 4 percent higher.   The  error  may  be due to
an increase in the particle size  and/or multiple  scattering
effects at low transmittance.   In either  case,  the error
would probably be reduced by employing a  smaller  receiving
angle .

-------
                      SECTION II

                    RECOMMENDATIONS
This program was  limited  to  field tests performed  at  one
facility.  It was  not  within the scope of the program to
extensively evaluate  optical properties of fly  ash under
varying size distribution and concentration as  related to
the opacity measurement.   It is recommended, however, that
such testing be  carried out.

As found during  tests,  the measured value of in-stack
opacity depends  strongly  on  the transmissometer  illumina-
tion and viewing  angles.   It may also be dependent on the
light beam diameter  used  for the aerosol illumination.
Tests, or at least theoretical calculations, should be
carried out to evaluate the  influence of this common  design
parameter.

-------
                       SECTION III

                      INTRODUCTION
Since the  turn  of  this century, a number of  theorists have
concerned  themselves with fundamental studies  to  explain,
predict or  calculate the extinction and scattering of a
beam of light passing through a suspension of  fine particles
or intercepting a  single particle.  Coincident with such
studies has  been the development of methods, techniques
and instruments to measure the magnitude of  such  phenomena.

As a result,  it is now possible to specify with reasonable
certainty  the optical conditions for extinction and scatter-
ing measurements in systems of particles of  undetermined
shape, size  and composition, and to anticipate what quan-
titative information about the particles can be deduced
from such  measurements.  Hodkinson1 has clearly and con-
cisely presented a review and critique of  the  noted activity,
described  methods  of applying the evolved  principles to
achieve suitable measurements and quantitatively  inter-
preted results  of  his and fellow investigators.  Conner
and Hodkinson2  have discussed in detail the  optical prop-
erties and  visual  effects of smoke-stack plumes.

Properly measured  and interpreted optical  properties of
plumes can  be related to the quantity and  quality of
material contained in a plume.  Capable of being  inter-
preted in  such  a manner, a measure of the  optical proper-
ties can be  used to indicate the effectiveness of system
processes,  applied control equipment and malfunction in
controls .

One of the  basic plume optical properties  is its  optical
transmi ttance which is defined as the ratio  of light flux
which reaches a light sensitive device  (eye, photocell,
etc.) when .the  flux from the source passes through the
plume and  when  it  does not.  The basic equation is given
by Equation  (1) :
                  T =  — = exp (-naQt)
                       o

-------
where

      T =  transmittance

      F =  flux  of  transmitted light through plume

     F  =  maximum  flux of  light measured in absence
           of plume in  line of source and sensor

      n =  the number of  particles per unit volume
           of air in the  light path of length t_
           through  the  aerosol

      a =  the projected  area of one of these
           particles

      Q =  the particle extinction coefficient or
           efficiency factor defined as:
   _ total  flux  scattered  and absorbed by a particle
     flux geometrically  incident on the particle
The particle extinction  coefficient or extinction effi-
ciency factor Q depends  on  the particle refractive index
relative to the surrounding medium, its shape and its size
relative to the wavelength  usually expressed as a = rrd/A ,
where d is the particle  diameter and A is the wavelength
of light in the medium  surrounding it; a is termed a
particle size parameter.

Another property of  the  plume is its opacity which is a
measure of the light  flux  attenuated by the plume that can
be calculated from the  transmittance as per Equation  (2):
                        0  =  1-T                       (2)
The optical density  (OD)  of  the same plume is expressed as
the logarithm of the  transmittance or opacity as per
Equation  (3):
             OD = -log1QT  =  -Iog10(l-0)               (3)

-------
From the discussion  and results presented by  Conner and
Hodkinson,2  a  remote method of measuring plume  opacity by
means of contrasting targets is described.  Actually,  this
is a calculated  plume transmittance value  (_T  )  obtained by
Equation  (4J:                                P
                                                     (4)
Bs and Bs'  are  the  luminance measured by  focusing with a
telephotometer  on  a specific target  (.sky)  respectively
clear and through  the plume.  B^ and B^'  are  the luminance
obtained when  focusing on a second target  (hill)  respect-
ively clear  and through the plume.  Performed in this
manner, a defineable plume transmittance  or opacity is
obtained.   It  is this value upon which  the plume opacity is
judged and  compliance to most control regulations is based.
Therefore,  if  in-stack transmittance measurements are to be
of value for determining plume opacity, they  must be relat-
able to such plume  measurements.  A significant difference
between plume  and  in-stack opacity measurements may be
realized in  the fact that the effective path  length may be
different.   In-stack opacity readings can  be  reduced in
practice to  the plume opacity by use of Equation (5) :
                                                     (5!
where

     0  =  plume opacity

     0  =  in-stack opacity

     £  =  stack exit or plume diameter

     H  =  in-stack transmissometer path  length.

-------
                       SECTION IV

                       OBJECTIVES
A number of  in-stack transmissometers are  commercially
available  for  the  continuous monitoring  of in-stack gas
transmittance.   However, due to the absence of defined
operation  and  performance specifications,  there has been
little standardization in design of such instruments with
regard to  illuminating and receiving angle, wavelength of
illumination,  sensitivity and type of receiver, and relation-
ship of output to  plume opacity.  Environmental Protection
Agency  (EPA)  officials and users have questioned the
accuracy and meaning of the data generated by such an array
of uniquely-designed transmissometers.   It was, therefore,
the object of  this study to answer these questions by
developing an  experimental transmissometer with extensive
research flexibility and collect and interpret data from
such a unit  while  operating it on a coal-burning electric
power plant  stack.

The experimental transmissometer employed  was flexible
enough in  design and the effluent stack  gases were of
adequate quality to determine the following:

     1)    The  effect of transmitter and  viewing angles
           on the measurement of transmittance;

     2)    The  dependence of illuminating wavelength on
           the  measurement of transmittance; and

     3)    The  relationship between in-stack transmittance
           and  plume opacity as a function  of in-stack
           path length.

This report  contains a description of the  transmissometers
employed in  the study, the experimental  test program
applied, and the results achieved.

-------
                       SECTION V

                    INSTRUMENTATION
Two in-stack  transmissometers Ca reference and  an  experi-
mental model)  were  designed,  developed and employed  in  this
study to achieve  the  results  reported.  The reference  model
was designed  with small  fixed transmitter and receiver
angles and was  used to continuously monitor the  transmittance
of gases flowing  from  the boiler to the atmosphere.   The
experimental  model  was designed with sufficient  flexibility
to permit changes to  be  made  in the transmitter  and  viewing
angles and also to  accommodate various interference  filters.
All other equipment used in the course of the program
activity was  commercially available.
               Reference Transmissometer

The reference  transmissometer consists of  a  transmitter
and a receiver.   In  operation the transmitter  is  mounted
on one wall  of a  duct  or stack and is the  source  of illum-
ination.   The  receiver is mounted directly opposite the
transmitter  and functions to measure that  portion of the
radiant energy which is transmitted through  the  gas flow-
ing within the confines of the stack boundaries.

Schematics of  the reference transmitter  and  receiver are
shown in Figures  1 and 2, respectively.  The basic support
and housing  for each is a rigid 3-inch I.D.  aluminum tube,
flat-black anodized  to eliminate excessive light  reflection,
Within this  tube  and at defined separation distances, as
shown in the figures,  are located the proper lens, pin
holes and  photocell.  The support tube was of  sufficient
internal diameter so that all internal items can  be first
mounted within an independent holding fixture.   At two
points  (diametrically  opposite) and near the circumference
of each fixture,  provisions were made to accept  a quarter-
inch rod and locking set-screw.  Two quarter-inch rods were
then attached  to  the first fixture and serially  extended
through each succeeding fixture.  The distance  between
fixtures was readily measured and held in  this  position by
tightening the locking set-screws.  The  entire  optical
assembly was then positioned in the support  tube  as shown
in Figures 1 and  2.
                          11

-------
to
                                Lamp Voltage
                                Regulating
                                Photovoltaic Cell
                                                                 Pin Hole
                                                                 0.0485 in. dia.
 Smoke Stop
'0.5 in.  dia.— Flange
Tungsten Filament Lamp
IYODA - 8V5A
                                                                               Flushing
                                                                                Air
                   Figure 1.   Schematic an4  Basic  Dimensions of  the
                                Reference Transmisaometer  Transmitter.

-------
                                             Flushing Air
- Connector
Selenium Cell
 IR A15M
/-Pin  Hole
  0.04  in
  dia .
                                                    I
   f
   w
   ^,
                   \v
                       2 . 39
                    3. 36
                                         13. 5
                                                   Smoke  Stop
                                                   1.5  in .  dia
       Figure 2
                                Schematic  and  Basic Dimensions  of  the
                                Reference  Transmissome ter Receiver.
                                                                                 rfl
                                                                                 •H
                                                                                 Tl
                                                  Flange

-------
As shown  in  Figure  1,  the transmitter optical assembly
consists  of  a  Tiyoda  8V5A square-shaped filament  lamp,  two
lenses, and  a  pinhole  of 0.0485-inch diameter.  The  optical
system is protected against dust deposits by introducing
clean flushing air  between the first .lens and the  stack.  A
smoke stop is  an  integral part of the support tube which
terminates in  a flange used for attaching the assembly  to
the stack wall.

The optical  assembly  tube and the illumination  lamp  are
mounted to a rigid  frame to secure optic alignment during
instrument operation  or transportation.  This frame  also
encompasses  the electronic components of the instrument and
is entirely  enclosed  in an aluminum housing with  a removable
cover for the  convenience of electronic circuitry  adjust-
ment and maintenance.   Complete assembly of the reference
transmissometer transmitter is shown in Figure  3.

The optical  portion of the reference receiver is  designed
to be assembled in  the same manner as the transmitter.  The
receiver consists of  a single lens protected from  dust
deposits by  a  clean air ventilated smoke stop,  followed by
a pinhole and  a IR  A15M selenium cell surface.  The
electronic system of  the transmitter/receiver is  located
in the transmitter  housing and performs two functions:  it
provides a stabilized  voltage to the lamp; and  conditions
the photocell  output  signal.  The wiring diagram  for  this
device is included  as  Appendix B of this report.

The stabilized  lamp voltage power supply maintains the
Tiyoda 8V5A  lamp  at a  nominal 6.5 volts.  The lamp intensity
is continuously monitored and adjusted for fluctuations
and aging by a  secondary photocell which is part  of  a feed-
back circuitry  of the  power supply.

The electrical  signal-conditioning circuitry consists of a
temperature-stabilized solid-state amplifier.   The
electrical signal is  received at the conditioning  circuit
from the receiver through a shielded weatherproof  cable
and, after proper conditioning, is suitable for 0-10 milli-
volt recorder  operation.  Electrical jacks are  provided on
the transmitter housing for continuously recording the
conditioned  signal.
                           14

-------
          I

Figure 3.  Photograph of the Reference
           Transmitter Assembly.
                 15

-------
              Experimental Transmissometer

The  separated transmitter and receiver concept was  also
applied  to  the experimental transmissometer design.

Several  techniques  for varying the transmitter and
receiver  angle of  transmissometers were investigated  and
the  final design  selected is shown in Figure 4.   The
experimental  transmissometer contains a lamp  (Tiyoda  8V5A)
which  can be  moved  along the transmitter support  tube  axis,
and  the  light from  the lamp is collected by a 5-inch  dia-
meter  condenser lens.   The light beam angle for the  lamp
located  close to  the condenser is about 100°; while  for a
lamp distance of  about 6 inches from the condenser,  the
light  beam  is nearly parallel and of very high intfinsity.
The  aluminum  support tube of 3-inch I.D. and all  parts of
the  optical system  were flat-black anodized to eliminate
excessive light reflection.

The  sheath  air provided to this system simultaneously
protects the  condenser lens surfaces and cools the  lamp.
This arrangement has the advantage of increased sheath air
temperature which prevents water vapor condensation  in
front  of the  transmitter condenser.

Possible light beam intensity changes caused by the  lamp
aging  or the  lamp power supply fluctuations are sensed by
a small diameter photocell located close to the lamp.  The
photocell output  is used as a feedback to the lamp  power
supply automatic  control circuitry that maintains the
light  intensity constant.  The photocell temperature  is
stabilized  to within ±0.5°F by a small wire resistor  which
is located  in the photocell mount.  The temperature  is
sensed by a thermistor which operates a Rosemount Tempera-
ture Controller STO-21501 that functions to control  the
energy to the resistor.

Similarly to  the reference transmitter, the experimental
transmitter is also enclosed in an aluminum housing  and
provided with power and low voltage shielded cord sockets.
The transmitter lamp voltage can be conveniently  checked
on two outside terminals and adjusted to the voltage  desired
upon removing the housing cover.  The light beam  angle can
be adjusted and the angle determined on a scale readable
through a small opening on the side of the transmitter
housing.  The opening  is sealed with a removable  cover when
                           16

-------
Secondary
Sheath Air
Primary
Sheath
                         Spacers

                         -Condenser
                                               Tungsten
                                               !•'i lament
                                             i  Lamp
                                               Light Intensity
                                               Control Photocell
              Figure 4.   Schematic  of; th.e  Experimental
                           Transmissometer  Transmitter.

-------
in use.   As  calibrated, the  transmitted light angle  can  be
varied within near parallel  zero  degree angle up to  100
degrees.   The lamp power supply  located in the transmitter
aluminum  housing is of the same  design as for the  reference
transmitter;  however, for the  experimental transmitter,  a
nominal  lamp  voltage of 8.0  volts was employed.

The  final experimental transmissometer receiver design  is
shown in  Figure 5.  It contains  a double aspheric  objective
lens that collects the light from the illuminated  section
of the stack  at angles from  about 110 degrees down to  2
degrees.   The exact angle is dependent on the iris opening
that is positioned between the objective and the light
focusing  lens.   As a light sensing device, a photovoltaic
cell A15M is  used in connection  with low noise solid-state
signal pre-amplifier, which  increases the-total signal
amplification of the secondary amplifier located in  the
transmitter  housing by 10 and  100.  The photovoltaic cell
is temperature-stabilized by the  Rosemount Temperature
Controller STO-21501.  The objective lenses are protected
by flushing  air supplied behind  the  smoke stop of  the
receiver.

The  electronic  circuitry of  the  experimental transmissometer
is shown  in Appendix B.


                   Auxiliary  Equipment

A telephotometer, Spectra Brightness Spot Meter Model  SB
(0.5° type) by  Photo Research  Corporation, was used  to
determine  the plume opacity.   The technique used in  this
determination was that of contrasting targets.

A set of  calibrated interference  filters supplied  by Optic
Technology,  Inc. was used to evaluate spectral behavior
of the transmissometers.  The  specific major peak  trans-
mittance  for  each of five respective filters was 436,  486,
579 and 656 millimicron.

Two  sources of  clean air to  protect  the optical systems
were constructed.  Each source consists of a small blower
and an absolute filter located in a  metal box.  Electrical
heaters with  proportional thermostatic control could be
                           18

-------
                             '—Focusing
                              Lens
                                   Photocel1
                                                 Pre-Amplifier
Aspheric
Obj e ctive
                     Interference
                     Filter
                                                 Heater
                                                              Thermostat
Figure  5.   Schematic of the  Experimental
            Transmissometer  Receiver.

-------
installed  in each assembly  if  h,igh temperature  flushing
sheath  air were required.   Air flow rate  for  each source
was about  7 cfm.

Two Bausch & Lomb VOM->5  single channel strip  chart
recorders  were used for  simultaneous monitoring of the
experimental and  reference  transmissometer  signals.
                           20

-------
                      SECTION VI

        LABORATORY  EVALUATION OF INSTRUMENTS
Functions of  the  instruments were checked and the  trans-
missometers calibrated in the laboratory b'efore  collecting
any field experimental data.  The attention was  mainly
centered on the  calibration of the reference and experi-
mental transmissometer output when instruments were  set to
an optical path  equal to the diameter of the stack.   For
each transmissometer  the amplifier linearity was determined
and the experimental  transmitter and receiving angle scale
factors were  established.  The data collected during labora-
tory calibration  were used to calculate the accuracy and
reproducibility  of  both transmissometers.
               Reference Transmissometer

The reference  transmissometer of 1.5° fixed transmitter
angle was designed  and used throughout the experiments.  The
receiver component  was first operated and evaluated  in the
laboratory with  a  3°  receiving angle but later,  when in-
stack transmittance data showed a strong dependence  on the
magnitude of the receiver angle, the design changes  were
made to decrease the  angle of view down to about 0.8°.

To determine the instrument sensitivity the entrance of the
receiver was shielded and the amplifier output  read  at the
so-called "dark  current" condition.  The output  voltage as
measured with  a  Keithley solid-state electrometer was within
0.0002 to 0.0003 volts.  This value represents  the minimum
accuracy of the  photovoltaic cell amplfier readout.

Based on this  calibration the minimum detectable transmit-
tance can be calculated.  When the receiver and  transmitter
were optically aligned and the transmitter lamp  operated
at 6.5 volts,  the  receiver signal corresponding  to clean air
conditions and light  path of 10 feet was 0.24 volts  and
for a 30-foot  distance, 0.0357 volts.  Supposing the ampli-
fier output reading of 0.002 volts has an accuracy of
±0.0002 volts, that is +10 percent, then the transmittance
determined with  the same accuracy would be:
                          21

-------
          0.002  x  100
     T =  	  0357	  =      percent for 30 ft distance

and

     T =   '  o  2400	  ~  0.83 percent for 10 ft distance.

The accuracy in measurement improves rapidly with increas-
ing trans-mittance and since the  transmittance values below
twenty percent  are  not  anticipated or considered allowable
during normal  stack emission rates, the performance is
considered to  be  acceptable.

The voltage  stability of  the photocell amplifier was deter-
mined by  step-wise  varying  the  amplifier input line voltage
from 115  to  100 volts.  No  detectable change in amplifier
output was detected.  The drift  of the amplifier during
startup was  less  than Q.5 percent of the output reading and
was stable after  one  hour of amplifier operation.

Similarly, the  lamp power supply was found to be stabilized
after one hour  of operation.  To simulate field operation
and noise pickup, the transmitter and receiver were sep-
arated a  distance of  9  feet, the light beam properly
aligned,  and the  photocell  amplifier output measured while
decreasing the  input  line voltage of the power supply to
87 percent of  its initial value.  The corresponding change
in the amplifier  output was less than 0,4 percent.  This
represented  a  combined  performance of the photocell
amplifier and  light source  stability.

The photocell  and amplifier circuit were selected and
designed  to provide for linearity between illumination and
amplifier output.   Two  linearity checks were performed to
determine the  true  relationship  between the photocell
illumination intensity  and  the  amplifier output voltage.
The first check was performed by utilizing a 15-watt high
intensity bulb  as a point source of light and causing this
light to  be  focused on  the  photocell from various known
distances and  at  various  levels  of intensity.  The ampli-
fier output was measured  and recorded at each distance
and intensity.  For a linear illumination-amplifier output
to exist, the  amplifier output  needs to be inversely
proportional to the square  of the lamp receiver distance
during constant light source intensity.  This was found to
be experimentally true.   When the lamp intensity was
                          22

-------
varied by varying  the voltage to the lamp, the  photocell
circuit linearity  characteristics were determined  for very
low photocell  illumination.   By this method  the  linearity
was shown to exist for amplifier output voltages which are
greater than 0.0002  volts.

The second photocell amplifier linearity readout check in
the laboratory consisted of  separating the transmitter and
receiver a distance  of approximately 12 feet (145  inches)
and inserting  neutral density filters in the transmitter
light beam.  The transmitter angle employed  in  this  check
was 1.5°, and  a receiver angle of view of  0.8°.  Six filters
having respective  neutral densities of 1.0,  0.7, 0.6, 0.4,
0.2 and 0.1 were serially inserted into the  light  beam
and the corresponding output voltage recorded.   For  each
respective neutral density filter the corresponding  filter
transmittance  is given as 10, 20, 25, 40,  63 and 80  percent.
The results of plotting known filter transmittance versus
reference amplifier  output in terms of volts is  given in
Figure 6 .

The angle of view  of the receiver was determined by
assembling the receiver in a horizontal position at  the
center of a circle 46 feet in diameter.  The distance from
the receiver to the  circumference was 23 feet and  a  move-
ment of 4.8 inches along the circumference represented a
one degree angle change.  This formation was used  in deter-
mining the angle of  view for a small 1.5-inch diameter high
intensity light source which was positioned  on  the 23-foot
arc and focused to the receiver.  The maximum photocell
output was noted and the light source was  moved in both
directions along the arc until a minimum photocell output
was recorded.   In  the first receiver design, a  movement  of
7.2 inches of  the  source either side of the  zero  angle was
required to produce  the minimum output reading.   This
corresponded to a  total movement of 14.4 inches or an angle
of view of 3 degrees.  When this same procedure was
employed for the receiver after it was redesigned  to reduce
the angle of view, it was found that total movement of
3.8 inches caused  the receiver to drop to  the minimum out-
put value.  This showed that the angle of  view was then
0.8 degrees .
                           23

-------
    100
    0)
    u
    G

NJ   4J
£>   4J
    •H
    6
    in
    c
Q)
-P
     50
                     lTr.an!smiut«r i
                         t     *  '  >
                     1    \-~-l-	~-|
                                                 0.1

                                Reference  Transmissometer Output, v
                                                                                                0. 2
        Figure  6.  Linearity of the  Reference Transmissometer  Readout as Determined
                     Using Neutral Density Light Filters.

-------
Prior to installation on the stack, the reference  trans-
missometer was  assembled and aligned in the  laboratory with
an optical path length of 145 inches to obtain  a  clean air
calibration  of  the  entire assembly.  This  calibration was
performed in  the laboratory after sunset to  obtain dark
room conditions.  The transmissometer operating character-
istics determined for the 3 degree receiver  angle  were
found to be  the following during this final  laboratory
calibration:

     1)   The collimated transmitter light beam focused
          on  the receiver was 2.5 inches in  diameter
          at  the receiver.

     2)   Operational stability was found  to be excellent
         'over  a line input voltage variation from 107
          to  116 volts.  At 104 volt input,  the lamp
          voltage decreases to 6.45 volts  from  the
          nominal 6.5 volt value.  The resultant  decrease
          in  lamp intensity and possible degradation in
          photocell amplifier circuit at this lower input
          voltage caused a decrease in photocell  output
          voltage of 0.007 volts or 1.8 percent of nominal
          value.  Increasing the line voltage to  125 volts
          caused the photocell output for  the same con-
          trolled lamp voltage to increase by 0.003 volts
          or  0.8 percent of nominal value.  The results
          of  this stability check with regard to  line
          voltage change is shown in Table 1.

     3)   The reference transmissometer startup and
          stabilizing period is ten to fifteen  minutes.
          The lamp  voltage and photocell output readings
          recorded  during the period of time from
          applying  a constant 116 volt input until stable
          readings  are recorded is shown in  Table 2.
   Table  1.   Stability of Reference  Transmissometer
              With Varying Line Voltages.
     Input
     [volts)
       104
       107
       116
       125
 Lamp
[volts)
 6 .45
 6.5
 6 .5
 6 .5
Electrometer  Reading
	(volts)	
       0 . 388
       0.395
       0 . 395
       0 .398
                           25

-------
    Table  2.   Stability of Reference  Transmissometer
               During Startup.
     Time  from Start    Lamp      Electrometer Reading
           (min)         (volts)          (volts)
              0            6.3             0 .382
              5            6.4             0 .390
             10            6.5             0. 395
             15            6.5             0.395
*
 Line voltage constant to 115 volts
Transmissometer  performance characteristics were also
established  for  the 0.8° receiver  angle  and with the
exception  that  the stable  (nominal  clean air)  photocell
amplifier  output voltage decreased  to  0.195 volts, the
stability  remained the same as for  the 3° receiver angle.
Nearly identical percent changes in  output were found for
variations in  line voltage in this  case.
              Experimental Transmissometer

Laboratory  evaluations of the experimental  transmissometer
were performed  to  determine system  stability,  angle of
transmittance,  angle of view, dark  current  photocell output,
and to calibrate  the output and determine the  accuracy of
the instrument.

A lamp power  supply voltage of 8.0  volts  is the desired
operational value  and this voltage  remains  constant with a
line input  voltage to the power supply  over a  range of
103 to 130  volts.   Over the same range  of voltages, the
amplifier output  signal is constant  for each fixed angle
of receiver and  transmitter.  This  range  of stability was
established by  repeatedly measuring  the lamp voltage and
recording the amplifier output for  various  incremental
settings of the  power supply input  voltage.

The various transmitter light beam  angles are  achieved by
adjusting the light source to a fixed distance from the
condenser lens.  The light source is mounted in a light
tube which  slides  internal to the aluminum  cylindrical
housing containing the condenser lens.   Therefore, it was
                           26

-------
necessary to  determine in the laboratory the  position of
the light tube  with respect to the lens.  Once  the  respec-
tive distance  and  position was established  for  each desired
light beam  angle,  the  outer surface of the  light  tube was
scribed and  labeled.   Establishment of the  light  tube for
each light  beam angle  (10° increments) was  achieved in the
laboratory  by  placing  the transmitter a fixed distance from
a plain dark  wall  and  focusing the positioned light source/
condenser beam onto the  wall.  The resultant  light  beam
diameter on  the wall  was measured and, knowing  the  distance
from the source to the wall, the sine/cosine  relationships
were employed  to determine the resultant transmitter light
beam angle.   In actuality, the required light beam  diameters
for each desired light beam angle were marked on  the wall
and the light  tube adjusted until that diameter beam was
obtained.   By  this method the transmitter was calibrated
for zero degree and in ten degree increments  up to  the
maximum beam  angle of  100°.

The experimental transmitter viewing angle  range  was estab-
lished in the  same manner as that described for the refer-
ence transmissometer.   By this method it was  found  that the
minimum viewing angle  of the experimental transmissometer
was 2° and  the maximum was 120°.

For various  viewing angles within this range, the viewing
angle adjustment mechanism was calibrated and scaled to
provide for  10° incremental changes.  Below 5°  a  special
pinhole plate  was  constructed and needs to  replace  the iris
to achieve  the 2°  angle  of view.

The "dark current" reading of the photocell,  preamplifier
and amplifier  circuitry  of the experimental transmissometer
was determined by  eliminating all light to  the  photocell
and measuring  the  resultant amplifier output  signal.
Readings were  obtained for all three preamplifier range
settings with  a Keithley 502 electrometer.   The results
are shown in  Table 3.
                           27

-------
           Table 3.  Dark Current  Signal  of
                     Experimental  Transmissometer
                     Receiver.
           Preamplifier     Dark  Current  Reading
              Range	     	(volts)	

              XI             -0 .00075±0 .00005
              X10            -0.0065  ±0.0005
              X100           -0.0660  iO.005
As shown  by  the table, for all preamplifier ranges the
"dark  current"  phenomena yields  a  negative voltage signal.
These  readings  are characteristic  of  most photocells and
the value  is  magnified by the increased preamplification.
However,  because of its small value  it  was found to be
difficult  to  eliminate electronically and achieve a true
base zero  for no illumination.   Therefore, for accurate
transmittance determinations the zero point offset must be
corrected  for arithmetically.  In  use,  the measured values
for any transmittance are increased  by  the amount of the
absolute  "dark  current" amplifier  reading specific for the
amplifier  range selected (XI, X10, X100).

The linearity of the photocell amplifier circuitry was
determined by the same two(point light  source and neutral
density filters)  techniques as described for the reference
transmissometer.  For the point  light source technique the
quantity  of  light reaching the photocell was varied by
changing  the  distance between source  and the receiver.  The
intensity  of  light to the receiver follows the inverse
square law as given by Equation  (6) :


                        E = KL~2                      (6)

where

     E =  amplifier output,  volts

     K =  constant

     L =  optical path between source  and receiver, feet.
                            28

-------
The corollary  to  Equation (6) with the data  obtained for
each of the  three preamplifier ranges was:
           — 2  3 8
     E = KL   '    for the XI range over values  of  E
         from  0.0014 to 0.028
           -2  19
     E = KL   '    for the X10 range over  values of E
         from  0.018  to 0.28
           -2  19
     E = KL   "    for the X100 range over values of E
         from  0.031  to 2.8.

The correlation  coefficient for each set of  linearity
measurements were within 0.995 to 0.999.   Even though the
linearity was  not defined at the  "dark current" values, it
is expected  from  the experience with the reference trans-
missometer that  acceptable linearity is  also achieved at
lower levels of  illumination.

In the case  of the neutral density filter technique to
determine photocell  and related circuitry linearity, tests
were performed for 2° angle of view and  for  transmitter
angles from  0  to  60°.  As shown in Figure 7, the  linearity
is good over the  range of transmitter angles from 10° to
60°.  At the  near 0-P transmitter  angle the generated
data was not of  sufficient quality to establish a good
linear relationship.  The reason  for this condition was
that the condenser of the light energy created a  discolor-
ation with the center and edges of the light beam at this
angle of focus and ND filters used appeared  to be color-
sensi tive.

A calibration  of  the experimental transmissometer was per-
formed to determine  the experimental transmitter light
intensity at the  optical distance for the transmitter and
receiver of  145  inches, and for receiver light angles of
2, 4, 5, 6.5,  8.5, and 16°.  Additionally, the transmitter
light angle  was  varied over the range of from 0° to 100° in
10 degree increments.  The receiver photocell amplifier
output corresponding to each transmitter light angle used
was measured by  a Keithley 502 electrometer  with no inter-
ference filters  inserted in the receiver optical system,
and the procedures repeated for 0.436, 0.486, 0.534 and
0.656 micron  light interference filters.
                           29

-------
        100
OJ
o
0)
o
a
td
    e
    w
    0)
    -p
    1-t
    •H
                               0.1

                               0.01
                                                 0.2

                                                 0.02
0.3

0.03
                              Experimental Transmissometer Output,  v
            Figure   7..
                     Linearity of the Experimental Transmissometer Readout as

                     Determined Using Neutral  Density Light  Filters.

-------
The illumination  lamp voltage of 8.0 volts was  set  and
used throughout the  experiments.  A near  linear relation-
ship between  the  amplifier output signal  and  the transmitter
light angle was found, except for the near 0° transmitter angle
For any given  receiver angle this relationship  can  be
expressed by  Equation (7):
where
                  E  = K x e



E = amplifier  output,  volts

K = constant

£ = transmitter  light  angle, degrees

b = exponent
                                                      (7)
The results  of  the experimental transmissometer  output
calibration  are tabulated and presented  in  Appendix A of
this report.  The  data are presented for  transmitter angles
from 0 to  60  degrees and receiver angles  2,  4  and 6.5
degrees.   For larger receiver angles,  the photocell output
signal for 6.5  degrees can be used because  it  was found
that the output is no longer dependent on the  receiver
angle magnitude above this value.

Using the  calibration data the accuracy  of  in-stack trans-
mittance determination can be calculated.   The accuracy is
generally  decreasing with increasing transmitter light beam
angle because the  light intensity as seen by  the photocell
also decreases.  This same tendency is also seen when
interference  filters are used in the receiver.

Table 4 contains the calculated accuracy  for  a given trans-
mittance when the  transmissometer is operated  at 100
degrees transmitter angle, 4 degrees receiver  angle and with
various interference filters in position.
                           31

-------
        Table 4.  Calculated Accuracy  of
                  Transmittance Measurements.
Illuminat ion
 of Filter
             rt  Accuracy as
Transmittance   Percent of
     (%)       Transmittance
 Accuracy of
Transmittance
White
0.436
0.486
0.534
0.656
Light
ym
ym
ym
ym
0 .8
30
10
5
5
±10
±15
±5
+ 2 .5
±2 .5
±o
±4
±0
±0
+ 0
.08
.5
.5
.13
.13
 Measured with 100° transmittance  angle,  4°  receiver angle
                          32

-------
                       SECTION VII

                       FIELD TESTS



            Field  Facilities and Conditions

The number eleven  pulverized coal-burning 100 MW  boiler and
associated stack  located at the Northern States Power
Company Highbridge Plant in St. Paul, Minnesota was  selected
and approved  as  the test site.  Tests performed in  January
of 1970 on this  boiler revealed the following data  as
measured downstream of the electrostatic precipitator :  gas
flow of 388,000  scfm;  stack gas temperature  about 200°F;
average dust  concentration 0.095 gms/ft^; precipitator
collection efficiency  approximately 93 percent.

As shown schematically in Figure 8, this stack is of
cylindrical steel  construction approximately  12 feet in
diameter and  292  feet  tall.  The measured internal  diameter
of the stack  is  145 inches and this dimension is  also the
effective light  path length with the transmissometers
located in position.  The stack protrudes through the
boiler room roof  which is 110 feet above the  ground. Within
the boiler room,  the boiler gases pass to a  split flow
electrostatic precipitator from which two separate  gas
streams exit  and  enter the stack just above  the roof line
through two diametrically opposite but identical  ducting
systems.  The experimental transmissometer was positioned
on the stack  50  feet above the roof.  The reference trans-
missometer was located 12 feet higher up.  The 12-foot
spacing between  transmissometers was determined to  be
sufficient to prevent  any interference between transmitters
during simultaneous operation.

Both transmitters  and  receivers were mounted  on the stack
wall using directionally adjustable ports and special
brackets welded  to the stack wall supporting  the  instru-
ments and locking them in the position desired, special
attention was given to locating the optical  system of the
experimental  transmitter and receiver close  to the  gas
stream flow boundaries to permit the evaluation of as large
transmitting  light angles as possible.
                           33

-------
      "Reference
    Tr a n s n i s s o rn e t e r
      Experimental
    Transmissoneter
                                L4 5 " I . D .
                                                      Boiler Room
                                                         Roof
Figure  8.   Experimental  Stack Dimensions
              and  Port  Locations.
                      34

-------
Two separate scaffoldings  attached to the stack provided
a base for the  transmitter and receiver operators.  The
transmissometers  were  electrically connected with suffi-
cient lengths of  shielded  cabling to the readout instru-
ments located in  a  shelter on  the boiler room roof.

During the field  evaluation the boiler was energized with
a mixture of approximately 10  percent Western Montana
powdered coal and 90 percent Illinois coal.  Boiler opera-
tion was steady during the periods of testing, but electro-
static precipitator rapping, performed at one-minute
intervals for about 15 seconds, caused some fluctuations
in the actual stack gas transmittance, as seen in Figures
10 and 11.  The plant  supervisor and engineer allowed  the
investigators to  vary  the  electrostatic precipitator's
high voltage power  supply  to obtain plume opacities within
limits of 4Q to 90  percent.  This was helpful in obtaining
data for a wide range  of the plume opacities, but evalua-
tion and interpretation of results became more difficult.
Th_is difficulty may have arisen from a change in the
particle collection efficiency of the precipitator for
various-sized particles, as the voltage was varied, causing
variations in the effluent particle size distribution.
Having no quantitative information about particle size
changes, no consideration  could be given to the changed
optical properties  of  particulates in relation to the  plume
transmittance.

The location which  provided the best view of the plume  and
permitted optimum opportunity  for obtaining contrasting
target plume opacity measurements was a river bank located
about 700 feet  southeast of stack number eleven.  From this
location the plume, as it  exited from the stack, was
readily visible.   A concrete stack under construction  at
the time extended above and to the left of stack number
eleven and served as the contrasting target for the  proper
plume opacity determinations.   A photograph of the two
stacks as taken from the plume viewing location is shown
in Figure 9.
                          35

-------
Figure 9.  Experimental Steel Stack* Viewed Against the
           Concrete Stack Used as the Contrasting Target
           for the Teleplxotometer Measurement of Plume
           Transmittance,
                          36

-------
        Field Optical  Alignment of Instruments

Each time the transmissometer was installed on  the  stack,
the following procedure  was followed to assure  proper
alignment of the  instruments:

     1)   The transmitter light beam angle was  set
          to the  minimum value to obtain a collimated
          light beam.

     2)   The covers  over the closed mounting ports
          were opened  and the transmitter was posi-
          tioned  in the  holder.

     3)   The transmitter was energized and aligned
          so that the  light beam was observed through
          the receiver port opening.  Alignment is
          achieved by  the observer on the receiver
          port side viewing the beam and directing  the
          individual  controlling the transmitter  to
          perform the  necessary adjustment until  the
          maximum light  intensity is visually realized.

     4)   The receiver is then positioned in  its  holder
          and slowly  adjusted up-down and left-right
          until the maximum photocell output  signal is
          establi shed.

     5)   When proper  alignment is achieved,  the
          transmitter  and receiver are physically
          locked  in position.

During the  course of  the field tests, each transmissometer
was removed and re-installed on the stack several times.
It was found that near identical alignment and  reproduci-
bility in results could  be achieved very rapidly  by follow-
ing the above-noted procedure.  The need for  repeated in-
stallation  was necessary because the demand for electrical
energy by the customers  of Northern States Power  Company
would not permit  a shut-down of the boiler for  periodic
clean air baseline photocell measurements.  Furthermore,
rigid routine lens and other component inspection schedules
were maintained to eliminate the possibility  of obtaining
erroneous information.  The air flushing system of the
transmissometers  performed properly for all components
except the  experimental  transmitter, where it was necessary
                           37

-------
to essentially expose the  condenser  lens  to  the  flue  gas
particles in order  to obtain maximum transmitter angles.
Due to the close proximity of  lens  and  contaminated
particles, this transmitter was  removed during  testing
activities every 60 minutes for  particulate  cleaning  if
needed and examination of  the  lens.
               Field Data  Procurement

With the transmissometers  properly  aligned  and  operating,
the photocell readout from each  transmissometer was
recorded on a Bausch & Lomb VOM-5  strip  chart recorder.
Examples of the resultant  chart  recordings  obtained  are
shown in Figures 10 and  11,   The reference  transmissometer
output recording was readily  related to  the plume opacity
and used to determine true in-stack transmittance.   The
experimental transmissometer  readings  were  related  to the
reference transmittance  to evaluate the  effect  of the
various modes of operation.

Chart recordings of Figures 10 and  11  show  only the  period
of opacity fluctuations  during the  time  electrostatic
precipitator rapping was occurring.  The  elapsed time is
approximately 15 to 20 seconds,  and each  of these periods
were followed by relatively steady  opacity  period of about
40 to 45 seconds duration.

For most experiments the sequence  of the  data taking was  to
scan the transmitter light beam  angles from 0 to 60  degrees
for a constant angle of  view  with  and  without interference
light filters in the receiver.   The angles  of view  of the
experimental receiver that were  tested were:  2, 3,  4, 5,  6.5,
8.5, 10, 16, 20, 30, 40, 50 and  60  degrees.  Transmitter
angles of near 0, 10, 20,  40  and 60 degrees were scanned.

Weather conditions permitting, telephotometer measurements
of the out-of-stack transmittance  were taken  using  the
technique of contrasting targets.

During some of the tests,  samples  were collected from the
stack through the experimental transmissometer  port  in order
to obtain some information on the  size distribution  of the
effluent particulates .   An isokinetic  probe,  incorporating
a filter holder containing  a  glass  fiber  absolute filter
                         38

-------
                     Run ;
                     Transmitter:
                     Receiver:
                     Filter:
lOb - 3/22/72
10°
40°
None
       EXPERIMENTAL  TRANSMISSOMETER
1
i "
1

i „
\ *
-j —
1

1 o
. t- - - -
' V
j £~~
1
I



-' .0
[










"




-




_-



-
	



















	

-


. ....


"

1
—
_....
-
—

}

L..






A


... j




' 1

:..


- --





. Ul
-



tr -



-


~ "


	
— .


L .


— _

_:_:"

--

~~—
tj


L
— -
— -


	
^_ 	 .

^1
	


-' -
— --'
	

T— :
t-rr-
—


h
LlJ
—
-— _•_
—
— -

_^r:_
r::_



^
^-1

.
3~

	
~^.~-T~~
	


-Z."


..

	 i



	
_._


i —
-

-
pf
f
ir.
__
L-1
....


r
_.-__
:-:-----
T"__
	






	
K--
	


UT
_^

— •;
— -
L _..

:.-.



-7
. -.
--


3 '.-.-
l_J
                Time,  seconds
         REFERENCE  TRANSMISSOMETER
Figure  1Q.   Example of the Experimental  Data
             Recorded in the Field Test.
                    39

-------
                         Run:          5a - 3/22/72
                         Transmitters  0°
                         Receiver:     3°
                         Filter:       0.656ym
            EXPERIMENTAL  TRANSMISSOMETER
                     Time, seconds
              REFERENCE TRANSMISSOMETER
         -3-
         2-
        -L-O

                     Time, seconds
Figure n.  Example  of the Experimental  Data
            Recorded in the Field Test.
                       40,

-------
(.Gelman Type E)  was  located in the gas flow, 4  feet  from the
stack wall.  The probe  was allowed to equilibrate  to  the
stack temperature and  a sample was drawn through  the  filter
where the particulate  collection occurred.  The dust  col-
lected was  evaluated for particle size using optical  and
electron microscopy.
         Evaluation of Field Transmittance  Data

The basic  parameters measured and recorded in the field for
further  evaluation were the reference  and  experimental trans-
missometer output signals and the telephotometer readings.
The reference and experimental transmissometer measurements
are used to calculate the in-stack  transmittance and to
describe the effects of operating the  experimental trans-
missometer with defined operational  parameters (i.e.,
transmitter and viewing angles).  The  reference in-stack
transmittance is calculated by use  of  Equation (8) :
                       T  = -
                        R   E
                             RT
                             RTO
where
        T   = calculated in-stack  reference transmittance,
             dimensionle ss
        RT
       RTO
measured in-stack  reference transmissometer
photocell output  signal,  volts

reference transmissometer photocell output
signal for  100  percent transmittance. Deter-
mined in the  laboratory for a defined optical
path of 145  inches and for the transmitter
angle of 1.5°.  For 3° receiver angle,  ERTO =
+0.240 volts;  for  0.8° receiver angle the
              RTO
                  value is +0.195  volts
The  experimental in-stack transmittance  is calculated by
means  of Equation (9):
                       T  =
                        E   E
                             ET
                                          (9)
                              ETO
                             41

-------
where
        T   =  experimental in-stack transmittance,
             dimensionless
       ET
       ETO
"dark current"  adjusted photocell
output, volts

"dark current"  adjusted calibration
photocell output  for laboratory determined
100 percent  transmittance at a given
transmissometer setting,  volts
EET was determined  by adding the established  "dark  current"
reading to  the  measured value.  The "dark current"  value  to
be added  is  dependent on the preamplifier scale used  to
obtain the  measurements.  The proper values as determined
in the laboratory  are given in Appendix A.
Different EETQ  values  were developed in the laboratory  for
each defined  receiver  and transmitter angle listed  and  for
each interference  filter employed.  The established  values
are given in  Tables  1  through 3 of Appendix A.

In spite of relatively good short term flue gas stability,
changes in transmi ttance were detected by the reference
transmissometer during prolonged periods of data taking.
Therefore, in order  to compare data from one readout period
to the next,  the experimental transmittance data was
normalized by applying the proper values in Equation (10) :
                 norm
                        — x T  = — x T
                        T     R   T    R
                                          (10)
where
       norm

         T
= normalized experimental  transmittance, %

= measured experimental  transmittance, %

= reference transmittance  corresponding to
                           42

-------
         T  =  average  reference transmittance for  the
               defined  array of measured experimental
               transmittance values being normalized

          T =  T  /T   relative in-stack transmittance.
               R E

By this method of normalization the normalized  experimental
transmittance  values are based on an average reference trans-
mittance for the defined array of measured experimental
transmittance  values,  and all values so normalized can be
compositely compared and analyzed.

The plume transmittance was calculated from Equation  (4) .
                           43

-------
                     SECTION VIII

                RESULTS  AND DISCUSSION


  Plume and Reference  In-Stack Transmittance Relationship

The average values  of  the reference in-stack and telephoto-
meter plume transmittance are presented in Table 5  and
plotted in Figure  12.   These results are for the reference
transmissometer operated with a 1.5° transmitter angle  and
a 3° angle of  view.  As seen from results in Figure  12,  a
plot of the experimental data provides a line which  has  a
slope just above unity.  With perfect agreement, the  data
should have fallen  on  the unit slope line.  It can  be seen
by the data points  that for transmittance close to  100  per-
cent this is  indeed realized.  For lower transmittances ,  the
measured in-stack  transmittance was slightly higher.   This
could be due  to the detection of scattered light by  the  3°
angle of view  of the in-stack transmissometer used  in these
experiments.
        Table 5.  Data on the Plume  Transmittance vs.
                  In-Stack Transmittance  Correlation.

NOTE:  Plume transmittance measured  by  Spectra SB 1/2
       Brightness Spot Meter by means of  contrasting
       targets technique; angle of view 0.5°.  In-stack
       transmittance measured at  1.5° transmitter angle
       and about 3° of the angle  of  view.

          Out-of-Stack            In-Stack
          Transmittance        Transmittance
                84                   87
                80                   81
                78                   76
                70                   74
                69                   68
                53                   56
                48                   49
                47                   53
                41                   46
                36                   38

 Linear regression of the  data:  Slope = 1.049; y intercept  =
 -5.316; correlation coefficient =  0.99.

                            45

-------
   100
o
u
c
(8
-P
4J
•H
e
CJ
C
flj
U
IT)
4J
O
I
4J
                               50
                                                        100
                    In-Stack Transmittance,  %
  Figure 12.
Transmittance Measurements by 3° Angle of
View and 1.5° Illumination Angle Trans-
missometer Inside  Stack  and 0.5° Angle of
View Telephotometer  Outside Stack,
                        46

-------
   In-Stack Transmittance  vs.  Experimental Transmitter
           and  Receiver  Angle  Characteristics

As previously discussed,  all of the experimental trans-
mittance data were  normalized  prior to studying the  effect
of the various  transmitter and receiver angles.

A plot of the normalized in-stack experimental transit-
tance versus the  operational receiver angles is shown  in
Figure 13 for various  transmitter angles.  The line  des-
cribing the transmittance  for  a near 0° transmitter  angle
is shown as a dashed  line  to indicate that it is based on
widely scattered  experimental  data due to the effect of
light beam discoloration caused by the condenser lens  at
this small angle.   However, the data clearly indicated the
same trend as shown for  the larger transmitter angles.

The data shown  in this plot demonstrates that the measured
transmittance is  strongly  dependent on the receiving and
transmitting angles.   The  contribution of light scattered
to the receiver from  particles inside and outside the  field
of the receiver view  is  given as the reason the experimental
transmittance is  always  higher than the corresponding
reference in-stack  transmittance.

From Figure 13  the  error due to any combination of  trans-
mitter and receiver angle  can be determined.  For example,
the error is about  9  percent for a receiver angle of 5°
and a 10° transmitter angle.

Data similar to that  presented in Figure 13 but for lower
values of average in-stack reference transmittance  were
reviewed during the course of the study and similar type
curves were generated.  For lower in-stack transmittance
the difference  between reference transmittance and  experi-
mental transmittance  was greater for all combinations  of
receiver and transmitter angles investigated.  This phen-
omena may be due to an increase in size of particles in  the
stack at the time of  the lower transmittance measurements.
The lower transmittance  in the stack was created by decreas-
ing the voltage in  the electrostatic precipitator  and,
consequently,  it is very likely that a difference  size
distribution of particles would penetrate the  precipitator.
A changed particle  size  distribution flowing through the
light beam would reflect a change in the relative  amount
of light scattered  to the receiver.  Ensor and Pilat  have
shown theoretically the  effect of particle size and trans-
missometer receiving  angles on transmissometer measurements.
                          47

-------
CO
      •o
      0)
      N
      (0
      6
      0)
      o
      c
      E
      ui
      c
      rfl
      ra
      4J
       0)
       cu
       X
       w
         1.2-
1.1
         1.0
0.9
          0.8  -•—
          0.7
                Range of In-Stack Reference
                Transmittance:  0.607-0.880
                Average In-Stack Reference
                Transmittance:   0.807
                Interference Filter:   None

                Run: 1, 6, 7, 8, 9,  12
                     of 4/19/72

                Reference Transmissometer
                Viewing Angle:  0.8°
                Illumination Angle:  1.5°

                Experimental Transmittance
                Normalized to 0.807  In-Stack
                Reference Transmittance
                      10
                       20
50
                                                             60
            Figure  13
                30        40
            Receiver Angle.  °
In-Stack Experimental Transmittance  Detected at
Various Transmitter  and  Receiver Angles of  the
Experimental  Transmissometer Normalized for True
Reference  In-Stack Transmittance of  0.807.

-------
These results  show  that  both the receiver and transmitter
angles of a transmissometer should be as small as practical
to restrict the  detection of scattered light and the
associated error.
    In-Stack Transmittance-Wavelength Characteristics

Tests with four  different light interference filters,
namely 0.436,  0.486,  0.579,  and 0.656 micron wavelength,
inserted in the  light path before the experimental receiver
photocell were performed to  define the measured trans-
mittance as a  function of wavelength.  The results of this
group of experiments  are seen in Figure 14 where the nor-
malized experimental  transmittance is plotted versus trans-
mitter angle for all  four filters evaluated.  These plots
show the dependence of the in-stack transmittance on wave-
length.  The 0.656 micron interference filter yields the
lowest in-stack  transmittance.   The transmittance increases
with each incremental decrease  in wavelength evaluated.
This result would indicate that spectral absorption effects
were present and/or the particulates were larger than about
one micron and sufficiently  monodisperse to observe oscil-
lation of extinction  efficiency with wavelength. Micro-
scopic analyses  of particulate  samples indicated the
number mean diameter  of around  1.4 micron and geometric
standard deviation of around 1.6.

The trend for  the difference between the measured experi-
mental transmittance  and reference or plume transmittance
to become greater for generally decreasing transmittance
was also seen  in this set of data.  Also, the value of  the
experimental measured transmittance is more dependent on
the magnitude  of the  receiver angle when filters are
employed.  This  phenomena results in generally  larger
experimental transmittance values to be measured with
relation to the  reference transmittance, and may be again
caused by differences in the effluent particle  size distri-
bution influencing the light scattering characteristics.

Secondly, the  rate of change in the difference  between
experimental and reference transmittance is greater for
increasing receiver and transmitter angles when filters  are
utilized.  This  phenomena may be caused by different  light
scattering characteristics of the effluent, due to  larger
fly ash particles present during the "interference  filters"
experiments.
                          49

-------
•d
0)
N
•rH
O
S3
ID
O
C
td
4J
jj
•H
s
0)
q
fd
rH
<0
•P
C
Hi
e
•H
1-1
X
w
    1.3
    1.2
1 . 1
    1 .0
    0.9
    0. 8 •
                     Jk;i trans a4..i;rahca:;D
                     f— ——-t— — ' -::!::;•:!:•:: ::;
Interference  Filters:

  x -0.4 36ym
  O - 0.486ym

  A - 0 . 579ym
  D - 0.65Gvim

Experimental  Receiver
Angle:  5°

Out-of-Stack  Transmittance:
0 .41-0.53

Reference  Transmissometer
Viewing  Angle:  3°
Illumination  Angle: 1.5

Experimental  Transmittance
Normalized  to 0.807 In-
Stack Reference
Transmittance
                                       30        40

                                 Transmitter Angle,
                                                                   60
              Figure 14.   Summary of  Interference  Filter  Tests.  Relationship
                             of  the  Experimental  In-Stack Transmittance  and
                             Transmitter Angle  for Four  Interference  Light Filters
                             at  5°  Receiver  Angle.

-------
Evaluation of the  transmittance data shows that the
closest agreement  between  in-stack and plume transmittance
measurements are achieved  when either white light or an
interference filter  between  0.486 and 0.579 micron is used
                          52

-------
                       SECTION IX

                    ACKNOWLEDGEMENTS
The work described in this report was performed  by
Environmental  Research Corporation  (ERC),  St.  Paul,
Minnesota, under EPA Contract No. 68-02-0309  by  a team
consisting of  Dr.  Carl M. Peterson, Dr. M.  Tomaides, Ben
Wahi, Keith  Rust,  Bob McKimmy, Len Graf and Roger Johnson.

The contract was administered under the direction of the
U. S. Environmental Protection Agency  (EPA),  with Mr.
William D. Conner as Project Officer.  The  authors wish to
thank Mr. Conner for his valuable contributions  and
guidance.

Also, they wish to express their appreciation to all
workers of Northern States Power Company  for their kind
assistance throughout the field experiments;  namely, to
Mr. Dave Williams, Patrick Gordon and Tom Anderson.
                            53

-------
                       SECTION X

                      REFERENCES
1.    Hodkinson, J.R.,  "The Optical Measurement of Aerosols"
     In: Aerosol  Science.   Edited by C.R. Davies, Chapter
     X, p. 287, Academic  Press,  New York, New York  1966.

2.    Conner, W.D. ,  and J.R.  Hodkinson  (1966) , "Observations
     on the Optical  Properties and Visual Effects of  Smoke
     Plumes", Environmental  Health Service, U.S. Govt.
     Printing Office,  PHS  Publication No.999-AP-30.

3.    Ensor, D.S.  and M.J.  Pilat  (1971),  "The Effect  of
     Particle Size  Distribution on Light Transmittance
     Measurement",  Am. Ind.  Hyg.  Assoc. J.,  32,  287-292.
                          55

-------
                      SECTION  XI

                      APPENDICES

                                                       Page

A - EXPERIMENTAL  DATA                                   59

    Figure A-l:   0.436  micron Interference Filter       64
                  Results

    Figure A-2 :   0.486  micron Interference Filter       65
                  Results

    Figure A-3:   0.579  micron Interference Filter       66
                  Results

    Figure A-4 :   0.656  micron Interference Filter       67
                  Results

    Figure A-5 :   No  Interference Filter Results         69

    Tables A-l  through  A-3:   Results of Experimental    61-
                  Transmissometer Calibration            63

    Tables A-4  through  A-15:  In-Stack Transmis-         70-
                  someter  Field Results                  81

B - NOTES TO  THE  REFERENCE AND EXPERIMENTAL             83
    TRANSMISSOMETER  OPERATION

    Figure B-l :   Reference Transmissometer Wiring       86
                  Diagram

    Figure B-2;   Experimental Transmissometer           87
                  Wiring Diagram
                          57

-------
                       APPENDIX A

                    EXPERIMENTAL DATA



   Results of  Experimental Transmissometer Calibration

Results of laboratory experimental transmi ssometer  calibra-
tion are presented.   When calibrating the instrument  in the
laboratory,  the  transmitter and receiver were optically
aligned and  firmly  attached in that position on two  sepa-
rate heavy laboratory benches, 145 inches apart.  The  output
of the receiver  photocell signal amplifier was monitored
with a high  input  impedance Keithley 602 Electrometer  and
by a Bausch  &  Lomb  strip chart recorder.  For constant
setting of the receiver angle of 2, 3, 4, 5 and 6.5  degrees,
the transmitter  angle was adjusted from 0 to 100  degrees,
and for each angle,  the amplifier output measured.   This
procedure was  repeated for four different interference
light filters  after  taking data for the transmi ssometer
operation without  any light filter in the line.   The  data
taken were corrected for "dark current" photocell operation
that was determined  by closing the receiver inlet and
reading the  amplifier output signal under this condition.
The "dark current"  was -0.00075 volt for "low" preamplifier
setting on the receiver panel, and -0.00650 volt  for
"medium" switch  position.

To correct the amplifier output readings, the following
equation was used:


                  ECOR - *EM-(ED.C.>
where
     E     =  corrected output
      COR
     E     =  amplifier output reading(was  negative
      M
             for low photocell illumination) ,  volt

     E     =  "dark current" amplifier output:
      D'C"   -0.00075 volt "low" preamplifier
             -0.00650 volt "medium" preamplifier
                           59

-------
The resulting corrected photocell  amplifier  outputs  are
tabulated for no interference  filter  in  Table  A-l;for
0.436 micron filter  in Table A-2;  and 0.656  micron  filter
in Table A-3; and the data  can readily be  used for  experi-
mental data evaluation or for  future  transmissometer
useage when the optical path length is 145 inches.
         Field Determination  of  "Dark  Current1
With the instruments optically  aligned  and  attached to the
stack, the  "dark current"  amplifier  output  was  determined
by closing  the receiver  inlet.   Because  this  value  depends
on the photocell temperature, this procedure  was  repeated
every hour  during data taking process.   This  detected
"dark current" amplifier output was  used to correct the
outputs determined for various  experimental transmissometer
operational conditions.  The procedure  used in  correcting
measured data for "dark  current" conditions was identical
to that described in the transmissometer calibration section
    In-Stack Transmittance-Wavelengtn  Characteristics

Tests with four different light  interference  filters, namely
0.436, 0.486, 0.579 and 0.656 micron wavelength,  inserted
in the light path before the experimental  receiver photocell
were performed to define the measured  transmittance as a
function 6f wavelength illumination or detection.

The results of this group of experiments  are  plotted in
Figures A-l, A-2, A-3 and A-4.   In these  figures  the
relative transmittance defined  in Equation (10)  is plotted
against the receiver angle as a  function  of the  various
transmitter angles.  As previously defined, the  larger the
relative transmittance number,  the lower  the  measured experi-
mental transmittance or normalized values.  It must be
noted that the true in-stack transmittance used  in calcul-
ating the relative transmittance was measured with the 3°
receiver angle reference transmissometer.  Based  on the
results described in Section VIII, all transmittance values
in Figures A-l, A-2, A-3 and A-4 would have been  about
5 percent lower if the 0.8 degree receiver angle  reference
transmitter had been used.  Such a correction would be
necessary if interference filters results  and no  interfer-
ence filter results are to be compared.
                          60

-------
Table A-l.
Results of  Experimental  Transmissometer
Calibration .



Transmitter
Angle
(degrees )
0
10
20
30
40
50
60
0
10
20
40
60
0
10
20
40
60



Receiver
Angle
(degrees )
2.0
11
tl
II >
n
"
M
4.0
M
It
M
It
6.5
II
II
"
II



Interference
Filter
(Vim)
None
M
n
it
ii
n
n
None
n
n
n
ti
None
ii
. n
n
n
*
Photocell
Amplifier
Output
(Corrected)
(v)
2.82075
0. 26375
0.09725
0. 06055
0 .04825
0. 04025
0.03455
3.35075
0.32075
0.12475
0.06375
0.04475
3.38075
0.33575
0.13275
0.06625
0. 04675
  Data  are  valid  for  "Low" pre-amplifier setting on the
  experimental  receiver cover.  For "Medium" setting,
   multiply values by  10.  Setting "High" is not
   recommended.

   Data  corrected for  the following "dark current"
   amplifier outputs:

      "Low"  pre-amplifier setting    = -0.00075 volt

      "Medium" pre-anplifier  setting = -O.OOG50 volt
                            61

-------
Table A-2.
Results of  Experimental  Transmissometer
Calibration.



Transmitter
Angle
(de grees)
0
10
20
30
40
50
60
0
10
20
40
60
0
10
20
40
60



Receiver
Angle
(degrees)
2.0
tt
tt
it
n
tt
it
4.0
ft
It
»
ft
6.5
It
It
n
n



Interference
Filter
(prn)
0.436
it
H
• ft
11
n
it
0.436
It
It
ft
11
0.436
it
n
n
M
it
Photocell
Amplifier
Output
(Corrected)
(v)
0.08975
0.00240
0.00075
0.00043
0.00032
0.00026
0 .00022
0.11275
0.00268
0.00088
0.00045
0.00033
0.11275
0.00268
0.00088
0.00045
0.00033
Data are valid for "Low"  pre-amplifier  setting on the
experimental receiver cover.   For  "Medium"  setting,
multiply values by 10.  Setting  "High"  is not
recommended.

Data corrected for the  following  "dark  current"
amplifier outputs:

  "Low" pre-amplifier setting    = -0.00075 volt

  "Medium" pre-amplifier  setting = -0.00650 volt
                        62

-------
Table A-3.   Results  of Experimental Transmissometer
              Calibration .

Transmitter
A a g 1 e
(degrees )
0
10
20
30
40
50
60
0
10
••» r\
40
60
0
10
20
40
60

Receiver
Angle
(degrees )
2.0
tl
11
II
II
II
11
4.0
II
11
II
II
6.5
ii
M
"
II

Interference
Filter
(Vira)
0.656
II
II
M
II
M
It
0.656
It
„
II
0 .656
•I
n
•I
n
*
Photocell
Amplifier
Output
(Corrected)
(v)
0.24375
0.00820
0 .00303
0.00180
0. 00140
0.00114
0.00097
0. 32075
0. 01095
0. 00200
0. 00138
0.33575
0. 01145
0.00440
0.00200
0. 00138
 Data are valid  for "Low" pre-amplifier setting on the
 experimental receiver cover.  For "Mediun" setting,
 multiply values by 10.  Setting "High" is not
 r ecominended .

 Data corrected  for the following "dark current"
 amplifier outputs:

   "Low" pre-amplifier setting    = -0.00075 volt

   "Medium" pre-anpli f i.er setting = -0.00650 volt
                           63

-------
  1.5
  1.4
                                                         60
                         Receiver angle,
Figure A-l.
Relationship of the  Relative Transmittance and
the Receiver and  Transmitter Angle for 0.436nm
Interference Filter.

-------
CTv
Ul
                          1.5
l-'
•
                          1.3  __
                       w
                      H
                      V   1.2
R
tcance
                      
                      •H
                      4J
                      (fl
                      .-I
                      <1)
                      C4
I-
•
                          1.0  -
E
W)
•2  o-'
H
O
•
CD
   0.7  -
O
•
CTi
                          0.5
                                       10
                           20        30        40
                            Receiver Am/le,  °
                                                                                        60
                    Figure A-2.   Relationship  of the  Relative Transmittance  and
                                   the  Receiver  and Transmitter Angle  for 0.486ym
                                   Interference  Filter.

-------
  w
 EH
 (U
 u
 4J
 4J
 •H
 'E
 W
 q
 m
 ti
 E-i

 0)
1.5 f



1.4



1.3



1.2



1.1



1.0



0.9



0.8



0.7



0.6



0.5
         Hti
                                      Tir an em it tor •'AngLe v:

                                                   :p:
                                                         28!
                                                 :J£r. armm ^ 11 an c_e
                                          fin
                   10
                       20        30

                         Receiver Angle,
                                               40
50
                                                             60
Figure  A-3.   Relationship of the  Relative  Transmittance and
               the Receiver and  Transmitter  Angle  for 0.579|im
               Interference Filter.

-------
 w
E-i
v,
 P
E-"
u
c
a
•H

4J

n)
     1.5
     1.4
     1.3  „.
     1.2  _:
     1.1  L-
     1.0
     0.9
     0. 8
     0.7  —
     0.6  -
     0.5
                                                    9.'.rM?J*S^


                                                     :-i:0:.,65

                                                    "~.'~~}~~'
                                          Trananittqr 'Andle::
FLlt^r':- ^ :-LO>p5G
             . . . . . "I . . 7
                    10
                              20         30         40


                                Receiver Angle,  °
                                                             50
                                                                        60
Figure  A-4,   Relationship  of the Relative Transmittance and

                the  Receiver  and  Transmitter Angle  for  0.656jJm

                Interference  Filter.

-------
The results can  be  summarized  as  follows:

1.   The plots show the  dependence of the  in-stack trans-
     mittance on wavelength  as  already described in
     Section VIII.

2.   The closest agreement between in-stack and plume
     transmittance  measurements are achieved when either
     white light or interference  filter between 0.486 and
     0.579 micron are  used.

3.   The trend for  the difference between  the measured
     experimental transmittance and reference or plume
     transmittance  to  become greater for generally de-
     creasing stack gas  transmittance was  also seen in this
     set of data.   The difference becomes  more evident for
     all cases when the  data of Figure 13  of the main
     report (no  filter)  is replotted in the same dimension-
     less transmittance  format  as Figures  A-l through A-4
     Cfilter data).  The  data  of  Figure 13 of the main
     report is presented  in  terms of the relative transmit-
     tance in Figure A-5.  The  results shown in Figures
     A-l through A-4 were obtained when the plume transmit-
     tance was generally  around 45 percent, whereas the
     data for Figure A-5  (no filter) was obtained at 60 to
     80 percent  transmittance.
              Tabulated  Experimental  Data

The representative data  collected  throughout the experiments
are presented for possible  further evaluation in the follow-
ing tables.  Tables A-4  through  A-7 contain the data
obtained when no interference  filters were used.  Tables
A-8 and A-9 were generated  when  a  0.436 micron filter was
used.  Data obtained while  using 0.486, 0.579 and 0.656
are respectively presented  in  Tables  A-10, A-ll, A-12,
A-13, A-14 and A-15.
                           68

-------
0)
O
c
0
in
O
rt
4J
•H
4J
   1.0
   0.9
   0. 8
   0. 7
   0.6
                10
        Figure A-5
   20        30        '0
      Receiver Angle, °
50
                                                               60
                                              Range of In-Stack
                                              Reference Tr ansmittance :
                                                0.607 - 0.880

                                              Reference Transmissometer
                                              Angle of View :
                                                0. 8°

                                              Interference Light
                                              Filters Used:
                                               None

                                               Run : 1, 6,  7,  8,
                                                   of 4/19/72
                                                                                     9,12
The  Receiver and  Transmitter Angle  Influence on  the Relative
In-Stack Transirjittance Expressed as  a  Ratio of the
Reference In-Stack Transmittance and of the Transmittance
Measured by the Experimental Transmissometer for  Given
Receiver and Transmitter Angles.

-------
Table  A-4 .   In-Stack Transmissometer  Field  Results.

c
3
K
1
1
1
1
1
1
1
1
7
7
7
7
7
7
7
7
7
c
tanco
Re f erenc
Transmit
2
0.612
0.607
0.744
0.723
0. 744
0.744
0. 752
0.764
0. 744
0. 764
0. 793
0.785
0. 773
0.855
0. 626
0.806
r*
1 U
41 O
C U
L'xpe rime
Tran smi t
Angle, c
3
10
10
20
20
40
60
60
10
10
10
20
20
2C
40
60
60
fH
4
4J
C
Exper ime
Rcccivei
Anglo, c
4
2
2
2
2
2
2
2
4
4
4
4
4
4 !
4
4
4
a
u
c E
U Zl
Interfei
Filter,
5
N *
N
N
N
N
N
N
::
;;
ti
N'
N1
N
H
N
N
«
.w
c
Experinu
Receive]
Range
6
L *
• L
L-
L
L
L
£,
. _ Z-
L
L
T^
L
L
L
L
1,
sd Zero Con-
No Aerosol
*n t al Trans .
V
4J - E -
0 C -H iJ
e> o ^ a
^ •* V fit
tt *J a, -w
o -H x y
u a u o
7
0.26475
0.26475
0.09325
0.09325
0.04925
0.03555
0.03555
i,. ^217.
0.32175
0.32175
0.12575
0.12575
0.12575
0.06475
0.04575
0.04575
intal
.somoter
V
Exper inu
Transmi ;
Output ,
8
0.169
0.170
O.Q78S
0.078
0.0395
0.0286
0.0293
J.264
0.259
0.269
0.113
0.113
0.111
0.0637
0 .0443
0.0436
id
?ntal
T some ter
V
ii E ~< -
0 ~i E U
4) W W ^
U O C 2,
i4 O< 
-------
Table A-5.   In-Stack  Transmissometer  Field Results.


c
a.
1
6
6 •
6
6
6
e'
6
6
a
8
8
a
8
8
0
u
c
c
C -rf
V*
V
£K E<
2
0. 859
0. 826
0. 847
0.859
0.859
0.806
0.859
0.347
0.830
0.839
0. 826
0. 818
0.835
0.847
•H
•tt M
•p o
c *J
g -4
h (fl o
0 C -H
a ra ci«
W t* <
3
10
10
20
20
40
60
60
60
10
20
20
40
60
60
r+
fa
u
c
e V
V4 ~4 0*
O CJ .H
a o tp
H a <
4
6.5
6.5
6.5
6.5
6.5
6.5
6.5
6. ^
8.5
8. 5
8.5
8.5
8.5
8.5
V
o
c 6
« 3
CJ -
h 0
0 -^
4J iH
K Et<
5
N *
K
N
N
;;
N
N
N
„

:i
I!
N
N
rH
«
4J
C
£ a
u •* o
e c c-
c. o c
B K C
6
L *
L
L '
L
L
L
L
L
T
L
L
L
L
L
•d Zero Con-
No Aerosol
ntdl Trans.
V
*J - E -
O o V4 D
W ~1 O C.
k -> C. -U
CJ Q U O
7
0. 33875
0. 33675
0.13375
0.13375
0. 06725
0. 04775
0.04775
: "47~ 5
0.34175
0.13375
0. 1337S
0. G672S
0.04775
0.04775
ntal
s o m c t o r
V
*? J
ti 0 3
oca.
C. n *J
a H o
8
0. 317
0.303
0.134
0.137
0.0706
0.0476
0. 0514
0.05- .
0.322
0.135
0.137
O.C6S7
0.0514
0.0531
d
n tnl
some tor
V
U E -i -
O W M O
M 0 C £.
K 0, H 4J
u u t> o
9
0. 31875
0. 30975
0. 13575
0.13375
0 . 07235
0.04935
0.05315
0.05315
0. 32375
0.13675
0.13875
0.07G45
0.05315
0.05485
ntal
t ancc

w in
o c
. 0, 0
U H
10
0.948
0 . 920
1. 015
1 .037
1 . 078
1.033
1.113
1.113
0.947
1 . 022
1. 037
1 . 047
1.113
1.149
n Ic s s
t a n c c
•H U -*
C *J f.
owe
E 1 H
a M £•
11
0.908
0.898
O.S35
0.823
0.799
0.780
0. 772
0.761
0.877
0. 820
3.797
0.781
0.750
0.737
cd
ntal
. t a n c e
-< £ •-<
n v< u:
E i c
U 0. «
Z M H
12
0.889
O.S39
0.966
0 .975
1.010
1.035
1.045
1.060
0.920
0. 984
1.012
1.033
1. 076
1.095
    No filter; L - Low preaaplifier setting.
                            71

-------
Table  A-6.   In-Stack Transmissometer Field Results.













c
3
a
1
9
9
9
9
9
9
9
9
9









c
<

CJ ^J
c -H
14 n
a c
u ic
.
X
*







_i
10
44
C
0 U
6 ai
Expcr
Rccc i
Range
6
L*
L
L'
L
L
L
L
L
r

1 *
C «H 0)
O O C
on*
O K
out-.
I* O
U < rt

0 4J
•a z c >

o a
E -I -
-4 6 U
oca
& « *>
X w 3
a E- o
8
0. 315
0.316
0.327
0.139
0.0769
0.0721
0 . 07 5 1
,.0534
0.0527





ti

4J
•H O
« e
4J O
•O C « >
o o  e -H -
U --4 g 4J
a ^ « 3
^4 a c o.
>4 a « u
0 X U 3
CJ M H O
9
0. 31675
0.31775
0.32875
0.14075
0.07865
0.07385
0. 07685
0.05515
0.05445







0
r* O
ti C
4J 
-------
Table A-7.   In-stack  Transmissometer  Field Results.
















«
1
1 2
12
12
12
. 12
12
12

12
12
12
12











o a

a £

o c

a H
2
0. 806
0. 797
0.335
0. 806
0.863
0.818
0 .347

0.368
0 .880
0.797
0 . 314








i-4
« U
AJ 1)
e a
£ ~<
-H E -
>w w O
o e -.
X V4 C
K r~ <
3
10
10
10
20
20
20
40

40
40
60
60








•_<
a
AJ
c
E o
-« > •
Vj -^ O
c c ^
X O C
a c <
4
60
60
60
60
60
60
60

60
60
60
60








o
u
C £
a 3.
o •
:
;;
N
;;
;;

N
N
K

i







^H
*f
4J
C
E 0
•x >
U -H a
l i r-
X O ^
w c «
6
^ *
L
L
L
L
L
L


^
L


I
C -< K
O O C
U n r;
0 H
0 U H
V4 C
0 < -1
N 0
0 u
•= z = >
** » e »
U C -H -P
C 0 V4 3
^ -H C Cj
0-4X3
U O tJ O
7
0.33675
0.33675
0. 33675
0.13075
0.13075
0. 13075
0.06725

0 .06725
0. 06725
0.04795
0.04795





w
o
AJ
•H O
n E
AJ O
C 0 >
E -i -
-< E -U
V CT 3
ace.
X V4 S
K E- t<
3
0.333
0.322
0.332
3.138
0.155
0.133
0.0757

O.OS16
0 . C S 1 4
0.0542
0 . 0548





ki
o
AJ
^ C
0 E
JJ 0
•3 C n >
AJ £ ^1 -
U -I E AJ
'j u n 3
^ c c £.
0 X ». 3
'J t: '• O
9
0. 33975
0. 32375
0.33375
0.13975
0. 15675
0.14075
0.0"'745

0.08335
0.08315
0.05595
0.05655







o
^ u

-------
Table
                                   In-Stack Transmissoirjeter Fieia  Results.
*»

c
3
p£
1
2a
2b
2c
2d
2s
14a
14b
14e
14d
14u
15a
ISb
15c
15d
ISe
ence
ni ttance
v c
WJ <0
u fc
a. H
2
0. 562
0. 562
0. 562
0.562
0. 562
0. 5G2
0.547
0. 531
0. 541
0. 531
0. 531
0.531
0'. 531
0.531
0. 531
•Iment al
mltter
o
Q. « cr>
x M c
U £-1 <
3
0
10
20
40
60
0
10
20
40
60
0
10
20
40
60
-iraental
uver
'-, °
04 U 0*
X 01 C
H K <
4
3
3
3
3
3
5
5
5
' 5
5
10
10
10
10
10
-f erence
ar, \ia (
4J rt
C -H
M fcl
5
A*
A
A
A
A
A
A
t-\
i
A
A
A
A
A
A
-imental
Lver
i
C, 0 C
K O a
U K K
6
H*
M
M
M
M
M
H

H
M
H
M
M
M
M
>cted Zero
L tion , No
sol Experimental
3 . Output , v
u rj o c
h C rt «
o o o n
O O «S E"
7
0.4918
0.0410
0.242
0. 0202
0. 0193
1. 2689
0.0509
0 . 0 2 '/ x
0. 0216
0.0202
1 . 3049
0. 0509
0.0274
0.0218
0.0203
-iraental
sraissoraeter
It, V
C' C C,
O, 1C *J
x M a
U H O
3
0.223
0.00276
-0.00656
-0. 00823
-0. 00383
O.C96
0.0107
-0. 00387
-0.00529
-0. 00762
0.721
0.0106
-0. 00276
-0.00605
-0.00689
;cted
riraental
;missometer
Jt, V
V4 CJ C Q<
U ft rO +J
O X W 3
U K H O
9
0.247
0. 02676
0. 01744
0. 01577
0.01517
0.720
0.0347
0. 02013
C . 01771
0.01638
0. 745
0.0346
0.02124
O.C1795
0.01714
rinental
sntittancc
u c
0« 11
X Li
W E-i
10
0. 502
0. 639
0. 721
0.781
0. 786
0. 567
0.682
0.743
0.820
0. 811
0. 571
0. 680
0.775
0.823
0.844
is ionic ss
tack
smittance
Q
•-t e h
a M H
11
1.12
0.88
0. 78
0.74
0.71
0.99
0 .80
0. 715
0.66
0.65
0.93
0.78
0.685
0.64
0.63
ilized
riraental
saittance
E « c
U O, nj
0 X U
2 W [-
12















set ed
Tsionless
smittance
k V C
>•< E *
O ~* w
o Q t.
13
1 .064
0.836
0. 741
0. 703
0. 674
0. 940
0 . 7SO
0.681
0.627
0.617
0.833
0.741
0.650
0.608
0. 598
                 A ~ 0.436  po interference filter; M - Medium preaaplifier setting.

-------
                       Table  A-^9 .   In-^Stack Tra,nsmis,someter Field  Results
m

c
3
C£
1
IGa
ICb
IGc
16d
IGe
17a
17b
17c
17d
17e
ce
tt ance
Re f ere n
Transni
2
0.531
0. 502
0 . £31
0.531
C. 541
0.541
0.525
C. 531
0.553
0. 541
ental
tter
o
Experim
Transni
Angl e ,
3
0
10'
20
40
60
0
10
20
40
60
rH
«J
+J
c
HI U 0
Experir
Receive
Angle ,
4
20
20
20
20
10
40
40
40
40
40
rence
Urn
Intcrf e
Filter,
5
A *
A
A
A
A
7,
A
A
A
A
ental
r
Expe r ir
Receive
Range
6
M*
M
K
M
K
M
;,
.~1
M
M
ed Zero
on , No
Experimental
Output , v
U 4J O •
a •*< w m
^ TJ O 1^
u c u at
0 0 o ^
O O < H
.7
1 . 304 9
0.0509
0.0274
0 . 0 2 1 S
0.0203
1 . 3040
C . 0 5 »
0.02:<
0.0213
0.0203
ental
ssomoter
V
•H E 4J
^ tn 3
o c a*
c. a -J
X l< P
WHO
8
0. 740
0.0127
-0. 00260
-0. 00633
-0. 00657
0. 75C
^ . 31 I
-0. 0022S
-0. 00531
-0.00629
od
icn t al
sso meter
V
U -H E 4J
a ^ w 3
n o s a,
M n< n 4J
O X M D
O W H O
9
0.772
0. 0367
0 . 0214
0.01767
0. 0174 3
C.780
0. 036C
0.02172
0. 01869
0.01771
i e n '^ a 1
ttanco
-H E
14 ffl
CJ C
o< n
X M
W H
10
0.591
0.721
0. 781
0.810
0.359
0 . 558
0.719
0.793
0. 857
0.372
on 1 o f s
-k
L 1 1 a n c e
D i m e r. s 3
In-Stac
Trail SITU
11
0. 90
0. 78
0. 68
0.65
0.63
0.90
0.73
0.67
0 . o 4
0.62
o
~l U
1 C
•a 4J «
0 C JJ
S C) 4J
Norma 1
Experir
Transm
12










.ed
Lonless
L 1 1 nnce
o tr E
QJ c ta
U V C
^ t -9
0 -H V<
U Q H
13
0.855
0.741
0.646
0. G17
0. 598
0.355
0.693
0 . 6 3 C
0 . 608
0. 539
                    0.436 pm interference filter; K » Medium preaaplifier setting

-------
                      Table A-10,   In-Stack  Transmissometer Field  Results.
er\
r • • •










c
3
K
1
3a
3b
3c
3d
3e
' n-.
19h
I9c
19d
19e
20a
20b
20c
20d
20e










m
0 U
U 4J
c -x
0 E
M 01
a c
u v<
2
0.537
0.525
0. 541
0. 547
0.537
0.528
0.531
0. 547
0. 541
0.531
0. 537
0.531
0.525
0.531
|o.531










erinent
nsmitte
Ic, °
K \* C.
W t- <
3
0
10
20
40
60
0
10
20
40
60
0
10
20
40
60









id
erinent
eiver
le, °
x 
o
erf eren
ter, (Jra
e -i
5
B *
B
B
B
B
B
B
-
a
B
B
B
B
B
B









«
icriment
eiver
igc
X o to
a ft K
e
M*
M
M
H
M
M


M
M
M
M
K
U
M
rl
«
*J
c
«J *>

0 -H *
U 0 W 4J
•u a o 3
M D. C.
- X 4J
•a c w a
0 O O
*J -H .-<
o -u o •
o -H in 01
w -a o e
o o o v*
O U < H
7
2.8863
0. 0798
0.0356
0.0271
0. 0240
3. 3709
0.1C2f3
o . o ; 3 c
0. 0311
0,0269
3 .4692
0.1033
0.0436
0.031S
0.0273





n
to
4J
.-< o
a e
J 0
C ol >
U tt
E -I -
•H e *>
W 07 3
CJ C 0,
X ti 3
WHO
8
1. 27
0.0231
0.0010
-0. 00312
-0.00532
1.65

0. 00822
0.00074
-:. 00268
1 .S3
0. 043
0.00823
0.0012
-0.0022





\4
o
*j
r-t V
a e
U 0
•o e B >
0 U ul
4-1 E •* -
0 •* E V
&> k w 3
w o c a<
0 X K 3
O W H O
9
1,29
0. 0471
0.0250
0. 02088
0, 01868
1.67
0.0643
0.03222
0.02479
0.02132
1.85
0. 067
0.03223
0. 0252
0.0218







o
^ u
n a
U K
t: a
v u
E -H
^. E
V in
u c
x n
U H
10
0.448
0, 590
0. 702
0.770
0.778
0.497
0. 628
0.719
0.797
0. 792
0.511
0. 643
0.739
0. 792
0.79B







n «
H U
c c
K *>
O M "
'H U -H
B fl E
£ ** W
a) in c
•H C V<
D u E-
11
1. 12
0.89
0.77
0.71
0.69
1 .06
0.84
0.73
0.68
0. 67
1.01
0. 32
0.71
0.67
0.66







01
~< u
*o c
rmalized
perincnt
ansmitta
0 x u
2 W H
12






















0} 0)
0} U
HI C
•O C 4J
0) 0 4J
AJ -r< -H
U to 6
0) C K
k « C
^ E rt
O -1 H
O Q H
13
1.064
0.845
0.731
0.674
0. 655
1 .007
0. 793
0.693
0.646
0. 636
0.959
0.779
0.674
0.636
0.627
                   B - 0.486 )im interference filterj M - Helium preamplifier setting.

-------
   Table A-ll.   In-Stack  Transmissometer  Field Results.

















3
PC
1
21a
21b
21c
21d
21e
22a
.72b
22c
22d
22Q








u
c








0) U
Pi E"*
2
0.547
0.531
0. 547
0.531
0.531
0.469
0.469
0.469
0 . 469
0 .469














-HE*

01 C ^H

X VJ C
W H <
3
0
10
20
40
60
0
10
20
40
60








.-4
10

C
01 H O
E  *

01 01 ^t
cu u o*
X flJ C
w B; «
4
20
20
20
20
20
40
0
40
40
40








0>
o
C E
01 XL

01 ~

^ OJ
01 -U

C -1
M Ix
5
D«
B
B
I)
B
B
3
B
B
B








(H
a

c
01 k4
E 01
•** >
M TH 0)
01 O CF>
n. u c
X 4) rtj
U fti QJ
6
M*
M
M
M
M
M
M
.1
M
H

ti
iJ
B
01 >
E
O -H •
K 0 H a
01 2 01 3
N o, a,
» X -U
T3 C W 3
01 O O

0 4J O •

u -a o c
U C U 0
O O 01 fc
U U < H
7
3.4692
0.1033
0 . 0436
0.0318
0.0273
3 . 4692
1.1033
0.04.6
0 . 0313
0.0273





H
O

r-l 
Cl 01
E •* •
•H E iJ
U 01 3
O C &
Q* rt *J
X ^ 3
WHO
8
1.93
0.0454
0 . 0100
0.0015S
-0 .00099
1 . 69
"l . J 3 L
0.00563
-0 .00105
-0.0040





H
a
4J
rH U
al E
•U 0
•a c 01 >
01 0) u

0 -H E •"
01 M 0] 3
H 0 C 0.
M Q. id *J
O X t4 3
U W £-• O
9
1. 95
0 . 0694
0 .0340
0.02558
0. 02301
1 . 71
0.062
0. 02963
0.02295
0. 0200







0
^ u
a C
+J rt
C 4J
a; ^>
C -H
•H E
M (A
Cl C

X W
W H
10
0. 563
0.672
0.780
O.S04
0.843
0.494
0.600
0.679
C.722
0.733







01 0)
01 U
01 C
r~\ 1)
C *J
O ^ *1
•H U -H
ol rt 2
C -M W
a w c

a H H
11
0.97
0.79
0.70
0.66
0.63
0. 95
0 . 78
0.09
C .65
0 . 64







a
H O
"3 C
TJ -P fl
Cl C 4J

•H C "1
-1 ~< E

12

















0) o
01 U
01 C

tj C *J


o 01 e
a c a

n 6 fl
O -1 W
U O H
13
0.921
0.750
0.665
0.627
0 . 598
0. 902
0.741
0 . £ 5 5
0.617
0. 608

0.486 wm interference filter; M - Mediun preamplifier setting.

-------
Table A-12.   In-St^ck Transmissometer  Field Results.

















tt
1
4a
4b
4c
4d
4e
25a
25b
25c
25d
2So












ot E




o; H
2
0. 566
0. 566
0.573
0.573
0. 576
0.533
0. 533
0.533
0.533
0.533
" 1 ""•
2Sa I 0. 533
26b
26c
26d
26e

0.533
0.533
0.533
0.533
t








*J O
G P


•** s *




W H <
3
0
10
20
40
60
0
10
20
40
60
0
10
20
40
60







,_(
at
4J
C
OHO
C c;
•rt > -
U -H V
CJ  3


IM M
k 0»
c p
JJ .— (
C "^
M &.
5
C *
c
c
c
c
c
c
c
• c
c
c
c
c
c
c







r4
n
I*
c
U 14
E 01
-H >
v< -H a
a) a» tr
Q, L> C
X 0) 0
w o: n:
6
H *
M
M
M
M
M
M
>.•

M
K
M
H
M
H

X]
C
ffl >
E
0 -< »
K O i, P
0) 2 01 P
M C. Q.
- X p
•O C W 3
at o O
p -H rH
U AJ O •
u -H n m
14 t) O C
h s u a
0 O 01 ^
U U «t H
7
0. 3292
0.1285
0,0530
0.0345
0.0294
6.0439
0.1634
^.0^.--
0.0414
0 . 0344
6 . 220S
0. 1701
0 . 0683
0 . 0426
0.0350




i.
Q
p
f< (U
<0 E
p O
C 01 >
C) tfi
F. -I -
-HEP
^ 01 3
O C C*
3. « P
X U 3
WHO
8
0.121
Q .0572
0. 0145
0 .00265
-0.00052
2 .75
0. 033
0.0244
0.0076
O.OC35
2.99
0 .090
0.0273
0.0090
0.00422




u

^j
(H 01
a e
P 0
•a c « >
o a u
P E -i -

01 U bl 3
k o c a.
M Cl< rt AJ
0 X W 3
U U H O
9
0.145
0.0812
0. 0385
0 . 02666
0.02343
2 . 77
0.107
0 . 0484
0.031S
0.0275
3.014
0.114
0. 0513
0.0330
0.02322






0)
r-t U
13 C
4j a
C .p
Ot 4J
e -i
•H E
u n
41 G
C, TO
X k(
U t.
10
0.440
0.632
0. 726
0.773
0. 799
0.459
0.635
0.720
0.763
0.799
0. 434
0.670
0. 751
0.775
0.806





en o
in u
o c
•H ffl
C u
O J^ AJ
-H U -H
0 « E
C -P W
O 10 C
S 1 *
•r4 C \,
Q HI fr"
11
1. 28
0. 39
0.79
L_2-'7
-------
                         Table  A-13 .    In-Stack  Transmissometer  Field  Results.
ID


C
3.
i
27a
27b
21:
27d
27o
2 C a
28h
2Bc
23d
23o
ce
ttance
aj K
^ (ft
<*4 (13
0) Vf
B; £-•
2
0. 540
0.533
0.533
0 . E 3 3
0.533
0.533
0.533
0.533
0.533
0.533
ental
tter
o
•H C -
ki w CJ
U fl C>
X U C
W £- <
3
0
10
:o
'10
60
0
) 0
so
•10
60
r-t
a
v>
c
V W O
•-I > -
W -H 01
o< u tp
X 11 C
u « <
4
20
20
20
20
20
40
,u
40
40
40
rence
ym
^ ti
w o
*j r-t
c -*
M r^
5
C *
c
c
c
c
c
c
.c
c
c
ental
r
•H >
M -rt 0)
a u c
X 41 fl
W « K
6
K*
]•:
K
M
(1
M
M
!t
M
M
ed Zero
on , No
Experimental
Ou t |
-------
                      Table A-^14 .   In-Stack  Tra.nsmissometer Field  Results.
CD
O
















K
1
5a
5b
5c
5d
So
30a
3 Ob
30c
30d
30e
31a
31b
31c
31d
31e




o
u
c

Cl 4J
U 4J

01 E
M 0)
oi c:


a. t>
2
0.531
0. 531
0.534
0.531
0.531
0.5-00
0.516
0.506
0. 516
0. 516
0.475
0.475
0.494
0.484
0. 500





PH
a to

C V
V JJ 0
E -i
••* E -
M 01 O
Cl C ^
O. flj D1
X k C

3
0
10
20
40
60
0
10
20
40
60
0
1C
20
40
60






<0
^j
c
0) U 0
E 01
•H > •
W -H 41
Ot 01 r-t
D. 0 0<
x a) c
U OS «C
4
3
3
3
3
3
5
5
b
5
S
10
10
10
10
10





d
u
c S
«1 3
^4
O "
*W V4
>j 01

U -rt O
Ct fij t7>
a u c
X OJ 13
•A a. a.
6
M*
M
M
M
M
H
M
.-
;i
H
M
M
M
M
H
r-4
C
0) >
E
O "H •
V4 O U 4J
01 SI <» 2
M a o.
- X AJ
•a c u a
01 O O
4J -r^ 1-4
U 4J O •
« ^i a a)
u -a o c
H C tl «
O 0 11 M
U U < H
7
4. 7159
0.1151
0. 0500
0.0339
0.0296
5.5069
0. 1499
u.Of,^
0.0407
0 . 0346
5.6679
0.1519
0.0641
0.0418
0.0353


^i

g 01
E -H -

M » 3
v c a
D. n) -u
X V< 3
WHO
8
1.54
0.035
0 . 00797
-0.00135
-0. 0029
1.88
0.0617
0.0159
0.00469
0. 00132
2.03
0. 0614
0.0194
0.00511
0.00215


K
0
4J
~f 0)
a E
U O
•a c 01 >
01 a a
*J E ^( •
u -rt E -u
QJ ^ W D
H « C 0.
W d <0 4J
O X V4 D
U M H 0
9
1 . 564
0.059
0.03197
0.02265
0.02110
1.904
0.0357
0.04090
0. 02369
0.02532
2.054
0.0854
0.0434
0.02911
0.02615




o
rH U
a c
*j in
C *J
0 *>
c ^
•H e

01 C
O< rt
X U
M H
10
0. 332
0.512
0 .639
0.668
0.713
0. 346
0 . 572
0.648
0.705
0.732
0.362
0. 562
0.677
0.696
0.741




01 a
01 U
a) c
fH id
C 4J
O A! 4J
•H O •*
U « 6
C -u m
01 Ul C
E i a
•rf C U
Q n H
11
1.6
1.03
0.835
0. 79
0 . 74
1.45
0 .90
0.78
0.73
0. 70
1 . 31
0.34
0.73
0.69
0.67




u
r* 0
n c
t3 *^ <3
O C •*>
H a *j
M g M
-i -H E
rt U ul-
E 0 C
M a, v
0 X U
2 W H
12



















n o
01 o
a c
•-< <
•O C *J

iJ M M
0 r, E
o c to
LJ O C
^ E fl
O -H K
o a t.
13
1 . 52
0.978
0. 793
0.75
0.703
1. 38
0.855
0.741
0.693
0. 665
1. 244
0.796
0.693
0.655
0.636
                    E • 0.656 Hm interference filter; H « Medium preamplifier setting.

-------
                    Table A-15.   In-Stack  Tran^missometer  Field Results.
CD














e
3.
i
32a
32b
32c
32d
32e>
33a
33b
33c
33d
33e








c


U 4J


01 C
<4-l (0
K I-
2
0. 500
0. 500
0. 500
0. 500
0.500
0. 462
0 . 500
0 . 500
0.516
0.484








«J 14


01 JJ 0
E -4


0. rt CT
U H <
3
0
10
20
40
60
0
10
20
40
60







r4
a
4J
C
01 M o
-4 > -
^ -H 01
01 01 -H
G, u en
X 01 C
(U Pi <
4
20
20
20
20
20
40
40
40
40
40







o
u
c e
01 ^


X4 01
01 4-»
4-1 «-4
C -^
M &-
5
E*
E
E
E
E
E
*
• E
E
E







r4
(0
4J
C

•H >
14 -H 01
C! 0) D1
Oi U C
X 01 d
u a: K
6
M*
H
M
K
M
M

rf
M
M

a
a
01 >
E

14 0 V4 4J
0 2 01 3
N a. a

•a c u 3
01 0 O
O *J O •
ai T4 w n
V4 TJ O C
K c ^ e
o o 01 M
u a < H
7
5 .6679
0.1519
0 . 0641
0. 0418
0.0353
5 .6679
0 .l^i-J
0 . 0641
0. 0418
0. 0353




'M
a
4J
^ O
•a E
4J O
G (n >
o> tn
-1 E 4J

o c c.
C« fl -U
X t4 3
W f- O
8
2.19
0. 0685
0. 0205
0 . 006S6
O.C0226
2.100
0.0704
0.0211
0. 00848
0. 00189




• u
01
4J
r-4 Cl
« E
•U O
•a c «] >

o -H E a

M 0) C p4
O X V4 3
U K f O
9
2.214
0. 0525
0. 0445
0. 03096
0.02626
2.124
0 . 0 & 4 4
0.0451
0. 03243
0.02589






01
M U
« c

C 4J
O 4J
•H E
tl 0)
0) C
X h
W t-
10
0. 390
0.610
0 . 694
0.741
0. 744
0.375
0.621
0.703
C. 777
0.733







rf -i E
rO tv W
E 01 C
O X 14
Z W H
12
















01 O
tn U
01 C

TT C *J
01 O 4-1
u n E

V4 01 C
0 -H V4
u a E*
13
1 . 216
0. 779
0.684
0 .636
0.536
1 . 1 6 H
0 . 7 i. 0
0.674
0.627
0.617

                  E - O.G56 Jim interference filter;
                                                 Mejiua preaaplificr setting.

-------
                       APPENDIX B

        NOTES TO  THE  REFERENCE AND EXPERIMENTAL
                TRANSMISSOMETER OPERATION
   Electrical  Hookup  and Alignment of the Instruments

Both transmissometers operate on 115 volts A.C., 60  Hz,
connected to  the  transmitter housing by a twist-type
electrical plug.   While operating, the transmitter  and
receiver must  be  interconnected by a shielded  cable  (about
15 feet long)  supplied with the instruments.   The amplifier
output signal  is  available from a ampheral plug  on  the side
of the transmitter housing and can be transmitted through
about 50 feet  of  shielded cable to any high  input impedance
voltmeter.  The  range of the readout for the  reference
transmissometer  should be from 0 to about 0.5  volt  depend-
ing on the optical distance of the transmitter and  receiver.
The range for  the experimental transmissometer should  be
from -0.001 to +7.0 volt.

After about 30 minutes warmup period, the power  supply
voltage for the  transmitter lamp must be adjusted to 8.0
volts on the  experimental transmitter, and 6.5 volts on the
reference transmitter.  The voltage should be  measured on
the lamp socket  of the reference transmitter  that is
accessible after  removing the cover.  After  attaching
measurement probes, the voltage must be checked  with the
cover closed  so  as not to influence the lamp  control photo-
electric circuit.  Two lamp voltage terminals  are provided
on an outside  panel of the experimental transmitter housing
and the cover  of  the experimental transmitter does  not have
to be opened  during the lamp voltage check.

The lamp voltage  can be adjusted in each instrument by
adjusting a trimmer that is labeled LAMP and which  is
accessible after  removing the transmitter housing  cover.

No other electrical alignment except the lamp voltage  is
required.
               Suggested Optical  Alignment

To align  the instrument on the stack,  the procedure des-
cribed  in Section III of this report  should be followed.

-------
                       Maintenance

No special maintenance of  the  reference  transmissometer is
required.  When  the  instrument is  operated  on a negative
pressure duct or  stack,  the  lens protection system is
sufficient to keep them clean  for  several  hundreds of hours
The only maintenance may be  the  lamp  replacement which can
be easily done after opening the reference  transmitter
cover.

The experimental  receiver  lens requires  checking and clean-
ing about every  80 hours of  operation and  the transmitter
lens must be checked and,  if necessary,  cleaned once an
hour.

When the instruments are to  be used on a positive pressure
duct or stack, a  high  pressure and possibly greater volume
of clean flushing air  to the optical  system must be
provided.

The lamp in the  experimental transmitter is replaced by
performing the following operations:

     1)   Remove  the instrument  housing  cover.

     2)   Disconnect the cable lead-ing from the lamp
          assembly to  the  electronic  component.

     3)   Slide  the  lamp assembly  out of the support
          tube housing.

     4)   Remove  the flushing/cooling air  base  fitting
          from the lamp  assembly.

     5)   Remove  the lamp  voltage  control  photocell
          located inside the lamp  assembly  tube.

     6)   Loosen  the three set-screws at the base of
          the lamp socket.

     7)   Turn and lift  out  lamp.

The new lamp may be installed  and  the instrument re-
assembled by following the described  lamp  removal process
in reverse.
                          84

-------
                   Electrical Diagrams

The electrical  diagrams  of the experimental and reference
transmissometers  shown in Figures B-l and B-2 are  enclosed
for ease in  trouble-shooting in the event that electrical
malfunction  occurs.

-------
                            /A/40O-3 (41
         /zov
CO
CTv
                                                                            EHVI*QNM£HTAL KESEAfTCH COKF>
                                                                             SCHEMATIC
                                                                                               £f?C-O/
                      Figure B-l .   Reference  Transmissometer  Wiring Diagram.

-------
(*m*0/J
BO9*T> I
1 •


fMfiOHHVn-4,. »eSfA»CH CMP
""*
'" '£-^'-TL
2>j rx- /".

— '••""



•'







'^fS^S4L»^Syfm^' "SB 100
     Figure  B-2 .   Experimental Transmissometer Wiring  Diagram

-------