2/3-5401
  IMPROVEMENT OF  OPTICAL EFFICIENCY
         OF  LUMINESCENCE  OF  A
     FLAME  PHOTOMETRIC  DETECTOR
            CONTRACT NO. EHSD 71-50
                 PREPARED FOR


         ENVIRONMENTAL PROTECTION AGENC
             RESEARCH TRIANGLE PARK
          DURHAM, NORTH CAROLINA 27711
             DECEMBER 1972
               FINAL REPORT
BAIRD-ATDMIG
GOVERNMENT SYSTEMS DIVISION


125 MIDDLESEX TURNPIKE, BEDFORD, MASSACHUSETTS O173O

-------
IMPROVEMENT OF OPTICAL EFFICIENCY OF LUMINESCENCE

           OF A FLAME PHOTOMETRIC DETECTOR
                   Contract No.  EHSD71-50
                         Prepared for

               Environmental Protection Agency
                   Research Triangle  Park
                Durham, North Carolina  27711
                        December 1 972
                         Final Report
Approved by:
                                L/L/ .
                        Arhur W. Hornig
                        Director of Research
                        Government Systems Division
                      Baird-Atomic,  Inc.
                    125 Middlesex Turnpike
                Bedford, Massachusetts 01730

-------
                          TABLE OF CONTENTS

Section                                                                 Page
1.      INTRODUCTION AND SUMMARY                                  1-1

2.      LABORATORY STUDIES                                          2-1
2.1    Standard Instrument Evaluation                                   2-1
2. 2    S  Emission Studies                                              2-1
2.3    Background Emission Studies                                     2-4
2.4    Choice  of Monochromator System                                  2-5
2.5    Multiple Slit Studies                                              2-6
2. 6    Background Subtract Mechanism                                  2-7
2.7    Electronic Modifications        -                                  2-7
2. 8    Laboratory Measurements                                        2-8
2.8.1  Modes of Operation                                               2-9
2.8.2  Single Slit Mode                                                   2-9
2.8.3  Multi-Slit Mode                                                   2-10
2.8.4  Multi-Slit Differential Mode                                       2-10
2. 9    Instrument Packaging and Controls                                2-11
2.9.1  Electronics                                                       2-12
2.9.2  Burner, Monochromator and Photomultip __.     	              2-12
2.9.3  Quartz  Plate Drive Mechanism                                    2-13
2.10    Delivery Demonstration at Research Triangle Park                2-13

3.      DISCUSSION AND RECOMMENDATIONS                           3-1
3.1    Recommendations                                                 3-2

4.      OPERATIONAL PROCEDURES                                    4-1
4. 1    Operating Controls                                               4-1
4.2    Operation Start-Up                                               4-1
4. 3    Recorder Input                                                   4-1
4.4    Calibration                                                       4-1
4. 5    Gas  Flow System                                                 4-2
4.6    Maintenance                                                      4-2
4. 7    Differential Mode                                                 4-2
                                                                              ii

-------
                          LIST OF ILLUSTRATIONS

Figure No.

     1         Vertical Spatial Scan

     2         Spatial Distribution of S  Emission (Clean Burner)

     3         Spatial Distribution of S  Emission (Dirty Burner)

     4         Spectral Distribution of S.  Emission  with 2 nm Bandwidth
                                        £
     5         Spectral Distribution of S  Emission/Background
               (Burner Center)

     6         Spectral Distribution of S  Emission  at Different
               Concentrations

     7         Spectral Distribution of Flame Background Emission

     8         Spatial Distribution of Flame Background (310 nm  line)

     9         S  Spectra at Exit Plane  of Monochromator for
               Different Entrance Slits

     10         Amplifier/Demodulator Schematic

     11         Spectral Distribution of S  Emission  with 3. 5 nm
               Bandwidth

     12         Spectral Distribution of S .Emission  with 5 nm
               Bandwidth

     13         Comparison of Multi-Slits and Single Slits

     14         SO  Monitoring with Multi-Slits in Differential  Mode

     15         Modified SO  Analyzer (Right Side)

     16         Modified SO  Analyzer (Left  Side)

     17         Modified SO  Analyzer (Top View)
                                                                                111

-------
1.      INTRODUCTION AND SUMMARY




       The principal objective of this contract was to increase the optical


efficiency of the  Melpar flame photometric detector for the detection of SO  .


Such increased efficiency should lead to higher sensitivity,  stability and


freedom from interference.  Two  routes were explored.  First,  the spatial


distribution of sulfur and background emissions were plotted to establish


regions of greatest signal/background.  Second, multi-slits were constructed


to  match the structured sulfur emission and allow background subtract.  The


spatial studies of sulfur and background emission  demonstrated that the opti-


mum  signal/background is obtained below  the geometric center of the stan-


dard burner.  The distribution suggested use of a  monochromator with


horizontal entrance  slit  rather  than a filter.  Unfortunately the distribution


is  not constant with time, being related to the buildup of deposits in the


burner.




       A study of the line emission of sulfur demonstrated that multiple


slits could be constructed which would both increase specificity for sulfur


and increase the usable signal.  A comparison of the multiple slits with a


single slit of the same  spectral  bandpass indicated an increase of about


three in  sensitivity.



       The differential mode, using the multi-slits, was not so successful


as hoped, for a variety of reasons.  First, the quartz plate introduces some


defocussing.  Together with the slightly oversized slits employed,  this re-


sulted in decreased  differential  signal.




       Defects in the burner and electronics of the standard instrument--


beyond the province of  this contract — limited the  results to a demonstration


of the successful implementation of the basic principles.   Using  multi-slits


in  the non-differential  mode the final s'ensitivity was about an order of -


magnitude greater than the comparison standard instrurrent (or about  three


times more sensitivity for SO ).
                             L*
                                                                            1-1

-------
2.      LABORATORY STUDIES





2. 1    Standard Instrument Evaluation





       The basic instrument supplied by EPA was a Melpar Model LL-1100-1B



SO. Analyzer.  As  received the instrument was very difficult to ignite and



demonstrated very  poor sensitivity.  The problem was traced to a leaking



sample chamber, resulting in a reduction in the hydrogen/air ratio.   While



the problem was temporarily  corrected by using a newer,  higher quality



O-ring and longer screws,  it  reoccurred during the entire project.





       The close coupling of the pump and the  sample/hydrogen sources re-



sulted in very visible "bouncing" of the balls  in the gas gauges.  A simple



expansion chamber, seven inches long and three inches in diameter, inserted



between the pump and the instrument,  increased flow stability and allowed



better setting of flow valves.  Subsequent work revealed that even then the



flows could not be accurately  reset using the  gauges provided.





       As the project progressed numerous leaks developed, particularly



around the igniter and flame ionization detector feed-throughs.   In each case



the problem was fixed in a temporary manner; however, it is clear that the



basic burner is not reliable.





2. 2    S. Emission Studies
       To modify the instrument for spatial studies of the combustion cham-



ber, we were able to reorientate the combustion chamber,  heated exhaust



manifold and inlet tubing without shortening any pieces  or changing heaters.



With the instrument cover removed, the combustion chamber was positioned



on the left side of the instrument,  oriented so it could be viewed by a



monochromator.   It was then a simple matter to restore the instrument to



its original  condition for comparison.





       The  burner was mounted on a  cross-slide allowing precise positioning



(±0.1mm) in frong of a 1mm   aperture; in turn, a quartz lens imaged the



aperture on the entrance slit of a Farrand monochromator.  The aperture
                                                                             2-1

-------
selected a small area of the flame region.   Various removable stops near


the lens allowed selection of angular field of view.   This arrangement allowed


versatile  spatial scanning of the combustion region at intervals of 1mm  and


was used  for both S.  and  flame background  monitoring.



       Using the NBS SO_ permeation tube  provided,  a concentration of
                         L*

approximately 1 ppm  was used for most studies.  First measurements  con-


sisted of horizontal and vertical spatial scans across the burner with the


monochromator set at an S  emission peak.  (At 1 ppm SO. flame background


is  no  problem. )  Figure 1  shows a typical set of raw data from a raster of


vertical scans over the combustion chamber. The circle in the upper right


hand corner represents the combustion chamber (dimensions X2).  The


square protruding from the bottom represents the flame shield; the nib  on


the left is the igniter; the  line  proceeding from the right hand wall represents


the (unused) flame  ionization detector.  The vertical scans are identified at


the top of the combustion chamber.  In the graphs the x-dimension indicates


the vertical distance  from bottom to top  (XZ).



       The data of Figure 1  indicate the  variation of S. emission.  Of parti-


cular interest are the dips in emission corresponding to the igniter and the


flame ionization detector.  Thus, Curve Z (left-most group) shows the  effect


of  the igniter while the effect of the larger FID element shows up in many


traces.  It is evident that  the greatest S. emission occurs below the horizon-
                                       L*

tal centerline and very close to the flame shield.



       Using data such as in Figure 1 and corresponding horizontal scans,


we had hoped to plot  contour graphs  of equal emission; however,  it was found


that by the time we completed  a set of scans we could no longer reproduce


the first scan in amplitude.  Thus,  the data were unsatisfactory.   The  non-


repeatability could have been due in  part to  the short-term stability of our


SO. source, but was  mostly due to actual changes within the burner.



       The contour graphs of  Figure Z were obtained by selecting a particular


emission  level  and scanning both x and y directions to find  all positions for
                                                                             Z-Z

-------
the given level.  In this way the time to complete a given contour was short,


and it was found the contours did form closed loops.  Longer term changes


would be reflected in relative error between contour  levels.



       Contour graphs such as these showed quite unequivocally that there


is a large variation  in S.  emission over the burner; and if Figure 2 were


representative, it would suggest using a horizontal slit across the two


peaks, or an off-center aperture for use with a filter.



       Unfortunately it was  found that  the pattern of flame emission varies,


within hours,  over a wide range,  due to turbulence in the flame and build-up


of a deposit on the inner walls of the combustion chamber.  After the chamber


has been disassembled and cleaned,  the emission pattern again shows a big


change.  In  Figure 3 we show contours obtained at a slightly later date


which show  a large asymmetry in emission.   In general, it was found that


the emission pattern depended strongly on the build-up of materials within


the burner.



       Other studies included the effect of angular  field of view on the emis-


sion pattern and on the changes associated with removal of the FID element


(which is not used).   While it is clear that protrusions  such as the FID ele-


ment or  the igniter do affect the patterns, the variability with buildup of


deposits  made  it  impossible to form a  clear picture of the effect of the


protrusions.



       In Figure 4 we indicate typical  S   emission  as a function of wave-


length using 2 nm spectral resolution on the monochromator, a 2mm slit


aperture at  the face of the burner and approximately  2 ppm SO..  In Figure 5


we  illustrate the  relative signal and background using 5 nm spectral resolu-


tion and  an  SO. concentration of approximately  0. 5 ppm.  In Figure 6, the
              b

spectral  resolution is also 5 nm but concentrations vary from approximately


1  to 0. 3 ppm.



       The  signal appearing at approximately 310 nm in both Figures  5  and


6 is the most prominent flame background,  probably  due to (OH).  Comparing
                                                                            2-3

-------
the center trace of Figure 6 and Figure 5 (identical spectral resolution and


SO. concentrations),  it is evident that the flame peak at 310 nm is much more
   L*

pronounced  relative to S. in Figure 5 than in Figure 6»   Figure 5 was taken
                       LJ

at the geometrical  center of the burner  and flame--the  region emphasized


by the standard Melpar unit.  Figure 6 was taken in a region exhibiting best


S  emission.  It is quite clear that the  signal/background is far from optimum


at the center of the burner.



2. 3    Background Emission Studies



       Figures 5 and 6 revealed the principal background signal at 310 nm.


In Figure 7  we show a typical background trace showing the 310 nm signal and


further emissions peaked at about 350 nm and very broadly at about 400 nm.


In this case the spectral bandwidth has been increased to just over 10 nm,


in order  to increase sensitivity.  From Figures 6 and 7 we find that  the ratio


of the highest S peak to the corresponding background  is about 800:1 for the
               L*

0. 5 ppm  SO. concentration.  Since the  S  signal depends on the square of the


SO. concentration, the background is equivalent to approximately 0. 02 ppm


SO .  While this is only an estimate, it establishes some idea of background


limitations.  This background does  not change much when our laboratory air


is filtered through  charcoal.



       The  data of Figure 7 were obtained using a cooled 1P28 photomultiplier


tube.  Using an uncooled  tube the electronic background interfered with flame


background determination.  Thus, for optimum sensitivity the photomultiplier


may have to be cooled,  although this requirement  may be mitigated by even-


tual optimization of the optical system.   In the standard Melpar unit  the


cathode of the photomultiplier is actually heated because of close proximity


to the burner  chamberl  This may well  account for the  instability and drift


observed during normal operation.
       Figure 8 depicts the spatial distribution of the  310 nm background


emission, using a measurement setup1 similar to that used for the S. con-


tours,  but with spectral bandwidth just greater than 10 nm.   Contrasting this
                                                                             2-4

-------
set of contours with the S  emission contours of Figure 2, it is  obvious that


it should be possible to select regions with optimum signal/background ratios.


In Figure 2 the maximum S  emission occurs at two points, both about 4mm
                          L*

below the geometric center of the burner chamber, near the corners of the


flame shield.  The flame background peak, on the other hand, is located


above the centerline of the burner.



       The original instrument has a 5. 5mm diameter aperture in front of


the burner.  This is indicated on both Figure 2 and Figure 8.  It is apparent


that it does not include the major  sulfur peaks, but is closer to  a background


peak.  Since the system  accepts a divergent beam (about 12°) broader areas


of the flame are actually viewed.



       Unfortunately a final signal/background plot (i.e.,  data  of Figure 2


divided by data of Figure 8)  cannot be made because the distributions change


within the burner.



       The chief background of interest is  actually not the 310 nm emission


(which occurs on the short wavelength side of the S. emission) but the 350 nm


peak and the broad peak  to longer wavelengths.  The 310 nm peak was used


for our measurements because  of its greater intensity;  however, an indepen-


dent measurement indicated that the 350 nm peak occurs in the  same spatial


position as the 310 nm peak.



2. 4    Choice of Monochromator  System



       As a result of the signal/background studies it was determined that


the chief emission regions for S  occurred near the edges of the flame shield,


somewhat below the centerline of the flame chamber; whereas the background


peaked above  the centerline  of the burner.  The optimum viewing  region


would then be below the centerline and include both S maximum emission


regions.   This could be accomplished using a monochromator with its


entrance slit horizontal and  in the plane of the emission maxima.   Or it


could be  accomplished by viewing only one maximum by means  of a lens and


an interference filter.  Both systems would allow background subtract;
                                                                             2-5

-------
however,  the filter system would be restricted to one line of S  emission



with resulting loss of sensitivity and specificity.  The monochromator was



to be preferred for the above reasons plus the versatility of tuning which



might allow application of the same instrument for other materials,  e. g, ,



phosphorus.





       With the project officer's  concurrence it was decided to incorporate



the  Farrand monochromator, used in the  studies of sections  2.3 and 2.4,



into the final instrument.





2. 5    Multiple Slit Studies





       The Farrand single monochromator has  a tuning range from 200-700 nm



and a dispersion of 11 nm/mm.  With exit slit removed the  exit aperture is



approximately  6 rnm,  corresponding to a  spectral width of  approximately



66 nm.  From Figure 4 it is evident that the six major S peaks could be



imaged simultaneously if the monochromator is set at 378 nm.  The average



distance between the major peaks  is 10 nm; however,  line  shapes are not



identical and it  is desirable to tailor a mask to fit the actual  emission lines



as observed by  a monochromator.





       The entrance slit will determine the spectral resolution of the mono-



chromator.  In  order to determine the optimum setting, pictures were taken



of the S  emission lines at the exit plane of the monochromator.  Entrance



slits of 2 nm,  3. 5 nm (fabricated specially) and 5 nm were employed for



the  data of Figure 9, the  monochromator  being set at 380 nm for  the 2 nm



slit and at 378  nm for  the wider slits.  SO  concentrations  were approximately



1 ppm for the 2  nm slit and 0. 75 ppm for the others.  From data such as these



it was  decided  to employ a 3. 5 nm entrance slit to compromise between



higher sensitivity and  adequate separation to allow measurements between



the  peaks  for background subtract.  Using separations measured  from the



photographs, 3.5  nm slits were etched in brass for use in the exit position



of the monochromator.
                                                                              2-6

-------
Z. 6    Background Subtract Mechanism



       In one mode  of operation the S.  peaks and valleys are to be imaged


alternately on the multi-slits in order to  subtract background and increase


specificity.   Several possible methods were investigated including oscilla-


tion of the multi-slits in the monochromator focal plane, rocking the grating


to move the emission image across the multi-slits, and rotation of a quartz


plate  in the  monochromator to move the image.  The last method was


selected for  simplicity of  mechanical operation and least interference with


monochromator tuning.



       A Z mm quartz -plate was mounted in the exit beam  of the monochroma-


tor, just before the  multi-slit.   The monochromator was tuned so that the


six strongest S. lines registered on the multi-slits  when the quartz plate
               L*

was oriented perpendicular to the  beam.  As the plate  is rotated in the


beam  the  S   lines are displaced until they are  not transmitted by the multi-


slits.   The quartz plate was driven by a 10 watt 115 volt a. c.  synchronous


motor, through a gear  train, to oscillate  at from 15-18 oscillations per


minute.  Oscillation  is caused by an eccentric groove  in  the gear face


which has a pin connected to a guided slide.  A pin in the slide connects to


a bell crank  on the quartz plate holder.  The bell crank has a slot for the


connecting pin  to allow for angular adjustment of the quartz plate.



Z. 7    Electronic Modifications



       While it was desired to use as much of the original electronics as


possible, it was not feasible to use the original package.   The original


amplifier had too long  a time constant as compared to the  chopping period.


A new electronic package  was developed which has  a wide-band pre-amplifier


stage, a synchronous demodulator,  an output filter with time  constants  of


0.1, 1. 0  and 10 seconds, and two stages of additional amplification.  The

                               -4     -10
full scale sensitivity is  from 10   to 10   amps.  Provision was made for


the addition of a log linear amplifier at a later date.
                                                                             2-7

-------
       The first stage  of the three-stage amplifier consists of an Analog


Devices 310J operational amplifier with associated feedback resistor to


make a transresistance circuit.  The output of this stage passes  through a


mechanical chopper synchronized with the optical chopper so that the de-


sired  S   signal is  detected when the quartz plate is perpendicular to the
       L*

optical beam.  The chopped  output is applied to the positive and re gative


inputs of an AD43J Analog Device through a time constant network.  The


AD43J is  used as a differential detector  with unity gain.  The differential


output is coupled to a Fairchild 741C amplifier which allows adjusting gain


by a factor of six for calibration  purposes.   The output of the 741C is  con-


nected to  a voltage divider,  giving both one volt and one millivolt outputs.


These outputs may be used to drive appropriate high impedance recorders


or a voltmeter.  Schematics  are  given in Fig.  10,



       During the course of our measurements we often  used laboratory


high voltage and low  voltage supplies in preference to the supply  incorporated


in the Melpar unit.  The built-in  high voltage supply had  a 120 cycle ripple


which contributed to  background.  However, for final  test we  returned to


the built-in supplies, since they sufficed to demonstrate  the operation of


the modified instrument.



2,8    Laboratory Measurements


       During laboratory measurements  prior to final assembly, after


final  assembly and at a demonstration at Research Triangle Park, many


difficulties occurred with the basic instrument.  These served not only to


delay performance checks, but to make detailed signal/noise  and sensitivity


measurements difficult.  The principal problems included recurring leaks


about gaskets, igniter  feedthrough and FID feedthrough,  shorting out of


block and exhaust heater control  (necessitating a manual system), and


somewhat unstable power supplies.  As a result the laboratory measurements


were often made with substitute amplifiers,  etc.  Nevertheless,  they demon-


strated the successful  operational mode  of the instrument.
                                                                            2-

-------
2. 8. 1  Modes of Operation





       The improved instrument will have  several possible modes of opera-



tion.  First the instrume nt may be used in  a mode similar to the original



instrument, using fixed single  slits.  These slits may be set  to match a parti-



cular line, or may be wide enough to encompass some or all  of the lines.



For extreme sensitivity  (and least specificity) the bandpass may be greater



than 60 nm and  include all six major lines.





       The instrument may also be used with the multi-slits  in the output to



give better specificity for S  emission.   The  entrance slit may be  matched to



the multi-slit (3. 5 nm) for best specificity, or  increased to 5 nm for  in-



creased  sensitivity.   This mode  is still non-differential.  The quartz plate



is adjusted perpendicular to the beam  and the oscillator is not used.





       Finally, using the multi-slits and a suitable entrance  slit,  the



oscillator may be used,  and the output  is the  smoothed difference between



measurements at S-,  peaks and valleys.   This last should offer the highest



specificity against interferents,  with some loss of sensitivity.





       The above discussion describes the application of the  instrument to



sulfur detection.  By tuning the monochromator to other wavelengths,  and



with suitable selection of slits,  the instrument is applicable to other  materials,



e.g. phosphorus.





Z.8.2  Single Slit Mode





       It is usually best to match input and output slits of the monochromator.



In Fig. 4 we displayed the spectral scan  of a 2  ppm concentration of SO



viewed with 2 nm slits.   It will be observed that the lines are well resolved,



with minima not far  from the base.  Fig. 11 contains  similar  data for 3. 5 nm



slits and a concentration of 0. 5 ppm.  The  sensitivity is clearly higher



while the resolution  remains fairly good.   The  minima approach the  base



less closely than in Fig. 4.   Finally, Fig.  12 shows similar data for 5 nm



slits.  The overall signal is greater, but there is  now a large unresolved



envelope.
                                                                              2-9

-------
       For greatest sensitivity  (at the loss of specificity) even larger


slits could be used.  Thus using 20 nm slits produces about four times


greater signal than 10 nm slits.



2.8.3  Multi-Slit Mode



       Using the line separation data of Fig. 9 and 3. 5 nm spectral width,


multi-slits were constructed for the six largest central lines  of S  emission.
                                                                L*

(Grooves  were first milled into  an aluminum block and photographed.


Slits were then etched in a nickel-plated copper of 0. 003 inch thickness. )



       The results were checked by comparing the throughput using  the


3.5 nm entrance slit and multi-slits (with the monochromator set to  pass


the six lines),  and the throughput using 3. 5 nm single entrance and exit


slits.  These results  are indicated in Fig. 13.  Here the lower trace shows


single-slit output while the upper curve  shows a scan through the maximum


for the multi-slit mode.  Gain is the same for both traces.  The increase


in sensitivity for monitoring six lines  over the  single largest  line (measured


by peak height) is about three, which is  somewhat lower than  if all lines


were imaged perfectly.   The increase is also about three if a peak to


valley  measurement is used.  We n ote that the  minima using  multi-slits


are not quite   so deep as hoped  for, indicating  the multi-slits are not


perfectly  designed.  As will be discussed  later, there is also a loss  of


image  definition  caused by the quartz plate.



       We have also obtained data for various larger entrance slits used in


conjunction with  the multi-slits.  The  output  increases more  or less line-


arly with input slit aperture; however,  specificity is lost rapidly (as •


measured by minima  when tuning with multi-slits in place).



2.8.4  Multi-Slit Differential Mode



       In this  mode the signal is sampled alternately when the S  peaks are
                                                               LJ

imaged onto the multi-slits,  and when they are shifted approximately 5  nm.


The sampling period is approximately 30°  for both peaks and valleys,  phasing


being accomplished by mechanical positioning of a reed switch.   From the
                                                                             2-10

-------
data accumulated it appeared best to use a 5 nm fixed entrance slit in com-


bination with the 3. 5 nm multi-slits.   (Spectral widths can be translated into


actual dimensions using the dispersion of 11 nm/mm. )



       Using filtered sample air,  the  output is  set to zero by adjusting the


reed switch.  S  emission is then  read as before.  Filtering is accomplished


after difference measurements.  Signal to noise is decreased not only because


of the use of narrow slits but because  of the short duty cycle.  (Information


is gathered for approximately one-sixth of a cycle. )



       Fig. 14 is a  tracing of cycling through a 10 ppb SO  sample and fil-

                                                                 - 8
tered laboratory air.  The  sensitivity  of the instrument is  set at 10   amperes


full scale and  a 10 second time constant is used.  The low-frequency  noise


is attributable  largely to the variation in phasing in the device, although


power supply instability is  also evident.



2. 9    Instrument Packaging and Controls



       With some rearrangement  of components it was possible to utilize


the original instrument case.  Side access holes had to be  added to incor-


porate the quartz plate drive mechanism and allow slits to be changed.  The


original front  panel was used, with relabeling of switches to conform to new


functions.   The final instrument is pictured in Figures 15-17.



       In Fig.  15 we see the instrument with right panel removed, showing


the new electronics sub-chassis as well as  the original front panel.   Figure


16 shows a view of  the left panel with quartz-plate oscillating mechanism and


slit access  holes.  A metal  dust cover for the drive assembly has been re-


moved.  Finally, Fig. 17 is a top view showing  the layout of principal com-


ponents.  The necessary rearrangements of components  will be discussed


below.
                                                                             2-11

-------
2.9.1  Electronics





       The new electronic package has the same overall dimensions as the



original.  The amplifier is mounted in a shielded subassembly, with three



access holes to adjust amplifier offset and gain.  The subassembly is mounted



directly to the rear of the front panel, utilizing the original cut-outs.  The



original scales apply to the internal calibration and to all amplifier settings



except the log  linear (which is  omitted).  The original two-position time



constant toggle switch has  been replaced with a four-position switch.   The



first three positions albw time constant selection of 0.1 seconds,  1. 0 seconds



and 10 seconds.  The fourth position allows conversion of the instrument from



the differential mode to a straight-through amplifier.  See Fig. 15 for  details.





2. 9. 2  Burner, Monochromator and Photomultiplier Detector





       The bulk of the  monochromator necessitated repositioning of several



components.  The monochromator is mounted  so that the  entrance slit is



horizontal and looking at the  lower center  region of the burner.  The burner



is mounted  directly onto the monochromator with an asbestos gasket for



thermal isolation.   The exit slit of the monochromator faces  upwards,



necessitating the acdition of a front-surface aluminized mirror to couple



the output to the original photomultiplier tube.  The latter is  now mounted



horizontally to fit into the original enclosure.  This  part of the layout can



be seen in Fig. 17.  Both entrance and exit slits can be  replaced through



openings  in the left  side  of the  cabinet (Fig.  16).  Standard slits of 2,  5, 10



and 20 nm together  with  a special 3. 5 nm  slit can be used in the entrance



position.  The exit position can utilize either the special multi-slit or  any



of the above.





       The monochromator dial indicator  has  been modified with a fine



tuning adjust located in the center of the calibrated dial to allow changes as



small as  0.1 nm.  With the fine adjust in its maximum counterclockwise



position the center wavelength  is as indicated on the large dial.  The fine



tuning adjust knob can be seen  in Fig. 17:  it is available only with the  top



to left side  cover  removed.




                                                                               2-12

-------
        The burner position was changed from rear center to front left cen-
ter to allow mounting on the monochromator.  Exhaust lines had to be
changed to accommodate the new burner location; however,  the location
of the sample control valve did not change,  and  the same rear panel  location
was used for the connection of the exhaust to the pump.
        The temperature control was relocated to allow space for the
monochromator and to  make adjustment of the temperature convenient.
        The sample air and hydrogen inputs  were moved from the rear panel to
the front left side,  reducing the length of the connecting hose to the flow indicators
to approximately 8  inches.  This reduction reduces recovery time  with
large sample  concentration changes and minimizes contamination of  lines.
2. 9. 3   Quartz Plate Drive Mechanism
        It was necessary to mount the motor and gears for the oscillating
quartz plate mechanism outside of the original case.  This can be seen in
Figs. 16 and 17.  The  unit is normally covered with a metal dust shield.
The oscillating mechanism can be  deactivated by means of a toggle switch
on the rear panel,  allowing use of  the instrument in the  original non-differen-
tial mode.
2.10    Delivery  Demonstration at  Research Triangle Park
        The demonstration at RTP  was not so  successful as we  had hoped due
to the recurrence of perennial problems with  the basic instrument.  A short
time  prior to the demonstration another leak developed in the igniter feed-
through.  Previous trouble with the automatic block/exhaust heater forced
use of a manual control.  Upon arrival  at RTP we found alignment of the
quartz plate, etc. ,  had been affected by shipment.   Further,  unknown  to us
at the time,  a further leak had  developed approximately equal in flow to
sample intake.
        The basic unit provided by  RTP for comparison was an  older model
which proved far more stable than the unit originally supplied to us. Initial
                                                                             2-13

-------
measurements revealed our unit was far down in sensitivity and signal to



noise.  After  intensive troubleshooting the instrument was demonstrated in



its several modes as discussed earlier.   The results of the  test may be sum-



marized as follows:





       1. The incorporation of a monochromator in the FPD was successful



in concept and execution.  The versatility in choice of bandpass,  single or



multi-slits, and tunability, was demonstrated.





       2.  The limitations of the  performance of the  revised instrument can



be traced to recurrent leak problems in the sampling system and a poor



burner geometry for coupling to the monochromator.





       3.  After troubleshooting on the spot,  the final sensitivity and



signal/noise  ratio of the revised instrument were slightly superior to those



of the comparison instrument.  Multi-slits  in the non-differential mode were



used for this comparison.





       4.  The differential mode  had considerably less sensitivity than the



non-differential.





       5.  Selectivity comparisons were inconclusive due to long sampling



lines  and uncontrolled  sampling me thods.





       Subsequent to the demonstration the large unknown leak was found



and repaired, increasing sensitivity and signal to noise by about an order



of magnitude.
                                                                          2-14

-------
3.     DISCUSSION AND RECOMMENDATIONS



       The principal objective  of this contract was to increase the optical


efficiency of the Melpar flame photometric detector for the detection of SO .
                                                                         C*

Such increased efficiency should lead to higher sensitivity, stability


and freedom from interference.  Two routes were explored.   First,  the


spatial distribution of sulfur and background emission were plotted to deter-


mine regions of greatest signal/background.  Second, multi-slits were


constructed to match the structured sulfur emission and allow background


subtract.



        An assumption made at  the beginning of this work was that the basic


instrument, including electronics and burner, was well designed and that


we would not attempt to improve on it.   Our experience has proven this


false,  since we have had to patch up innumerable leaks and make  do  with


an inadequate burner.



        The  spatial studies of sulfur and background emission demonstrated


that the optimum  signal/background is  obtained below the geometric  center


of the  standard burner.  The distribution suggested use of a  monochromator


with horizontal slit rather  than a filter.  Unfortunately the distribution


is not  constant with time, being related to the buildup of deposits  in  the


burner.



        While the  signal/background ratio is increased with the  monochromator,


the total light input is decreased considerably,  limiting any dramatic in-


crease in sensitivity.  The obvious next step is to redesign the  burner to


maximize coupling of signal to a monochromator slit.



        A study of the line emission of  sulfur demonstrated that multiple


slits could  be constructed which would both increase specificity for sulfur


and increase the usable signal. A comparison of the multiple slits with  a


single slit of same spectral width indicated an increase of about three in


sensitivity.  The  multi-slits fabricated at Baird-Atomic are not optimum;


however, they demonstrated the correctness of the principle.
                                                                              3-1

-------
        The differential mode, using the multi-slits, was not so successful



as hoped, for a variety of reasons.  First, the method chosen to shift the



spectrum across the multi-slits involved some difficulty.  The quartz plate



introduces  some defocussing.  Further, phasing, or balancing of peak



and valley signals, was not well enough controlled using reed switches,



resulting in some instability.  Finally, the greater bandwidth required



resulted in increased noise.  A further consideration involves choice of



slit widths.  For the non-differential application of multi-slits  a 5 nm en-



trance slit and 3.5 nm multi-slits were chosen as  a good compromise be-



tween high specificity and high throughput.  In the  differential mode it




would have been better to use narrower slits  since peak-valley  differences



are quite sensitive to spectral width and precision of registration.





        A plague of burner leaks and electronic problems in the basic  instru-



ment,  particularly in the latter  part of the study,  made it impossible  to



determine the increase in sensitivity afforded by the innovations introduced.



Using a 5 nm entrance slit and a 3. 5 nm multi-slits in the non-differential



mode, the demonstration showed a better  signal/noise than the  comparison



standard instrument. The test data were  not precise, but suggest  an  im-



provement of about a factor of three.   Since this measurement was  made



with an unknown leak present, we feel the improvement was  probably  well



over an order of magnitude.





        In conclusion, the work of this  contract has successfully demonstrated



the use of a monochromator and multi-slits to increase both the optical



efficiency and specificity for sulfur detection in a standard flame photometric



detector.





3. 1     Re commendations





        Based on the results of this  program the following  recommendations



are made to realize  fully the potential  of techniques developed:





        1.  Redesign  the burner to optimize coupling between sulfur  emission



and a monochromator slit.
                                                                            3-2

-------
       2.  Redesign the burner to improve flame stability and eliminate
leaks.
       3.  Evaluate improvement in signal/noise and establish minimum
detectable level of SO .
       4.  Investigate applicability of instrument to other materials,  e. g. ,
phosphorus,  arsenic and selenium.
                                                                              3-3

-------
4.
       OPERATIONAL PROCEDURES
4. 2
4.4
       a.  H  Flow Control
       Because of the similarity to the basic Melpar unit,  most ope rational


procedures are similar to those for the model LL-1100-1B SO  Analyzer.
                                                            L*

Reference should  be made to the Instruction Manual, especially Section III.



4.1    Operating  Controls



                                      Same


       b.  Sample Air  Flow Control   Same


       c.  Power Switch


       d.  Ignition Switch


       e.  Temperature Control


       f.  Range Selector


       g.  Calibrate Switch
       h.  Time Constant
       i.   Buckout


       j.   Thermocouples






       Operation Start-Up



       a.  Set-up Provisions





       b.  Operation
4. 3    Recorder Input
       Calibration
Same


Same


Same, but relocated


Same, but no LOG position


Same


Changed to rotary switch with time


constants of 0. 5 sec, 1 sec and 10  sec.


Two spare positions available for


other time constants.


Same


Same (but  temperature control is


manual using a Variac)








Same, except set  RANGE SELECTOR


switch to appropriate linear scale.


All items the same except linear


scale only.



Same, except omit reference to LOG


position.



Final calibration incomplete due to


various possible modes  of operation


and difficulties with basic instrument.
                                                                             4-1

-------
4. 5    Gas Flow System               Same (Section IV of Melpar manual)





4. 6    Maintenance                   Same (Section V of Melpar manual)





4. 7    Differential Mode



       The differential mode is in operation when the chopper toggle switch



on the rear of the instrument is in the on position.





       To shift from differential mode to straight operational mode, turn



chopper  switch  off.  Remove dust cover on chopper assembly and rotate



chopper  drive so that the upper reed  switch is in  a closed position.  The



detector output  is then connected directly to the operation amplifier for a



direct output to recorder.  The chopper switch in off position closes opera-



tional amplifier positive input to ground.
                                                                             4-2

-------

       FIGURE 1
VERTICAL SPATIAL SCAN

-------
BAIRO ATOMIC
                                                                     SPATIAL DISTRIBUTION OF S2 EMISSION
                                                                                (Clean Burner)
                                                                     Slits:  10-20/1 mm2 aperture
                                                                     Concentration:  ~1.0ppm
                                                                     Scale:  1 cm  = 1 mm

   IGNITER
        8

-------
                                                                                           BAIRD ATOMIC
      IGNITER
SPATIAL DISTRIBUTION OF $2 EMISSION
          (Dirty Burner)
Slits: 10-20/1 mm2 aperture
Concentration:  ~ 1.0 ppm
Scale: 1 cm  = 1 mm
BURNER
  SHIELD

-------
     BAIRD ATOMIC
                                                                      SPECTRAL DISTRIBUTION OF S- EMISSION
                                                                      Slits:  2-2/2 mm col. slit
                                                                      Concentration: ~2.0ppm
200
300
400                      500
  WAVELENGTH (nanometers)
600

-------
                              FIGURE 5

                    SPECTRAL DISTRIBUTION OF S,
                      EMISSION/BACKGROUND l
                          (Burner Center)
                    Slits:  5-5/2 mm col. slit
                    Concentration:  ~ 0.5 ppm
                                                              TOO
WAVELENGTH (nanometers)

-------
                                                                                          FIGURE 6

                                                                           SPECTRAL DISTRIBUTION OF S2 EMISSION
                                                                              AT DIFFERENT CONCENTRATIONS
                                                                           Slits: 5-5/2 mm col. slit
                                                                           Concentrations:  A ~ 1.0 ppm
                                                                                          B ~ 0.5 ppm
                                                                                          C ~ 0.33 ppm
200
300
400                    500
  WAVELENGTH (nanometers)
                                                                                            600
                                                                                             700

-------
                                                FIGURE 7

                                     SPECTRAL DISTRIBUTION OF FLAME
                                         BACKGROUND EMISSION
                                     Slits: 10-20
                                                                        700
WAVELENGTH ( nanometers  )

-------
  BAIRD ATOMIC
SPATIAL DISTRIBUTION OF FLAME
  BACKGROUND (310 nm line)
Slits: 10-20/1 mm2 aperture
Sample: - ambient air
Scale: 1 cm  = 1 mm
BURNER  SHIELD

-------
                            2 nm entrance slit,
                  1.0 ppm SO,-,  380 nm center wavelength
                          3.5 nm entrance slit,
                 0.75 ppm SO9, 378 nm center wavelength
                            5 nm entrance slit.
                 0.75 ppm SCX,  378 nm center wavelength         photos enlarged 3X
Figure 9.  So spectra at exit plane of monochromator for different entrance slits

-------
           RANGE




           IOOOM
f 15
1


8UCKOUT IOOM
* _L*

r 1- IMF







PM i




-15
1
4.99K
IOK

IK
IM


IOOK
^. 1 M 	 ,
IOM
^lOOM^
^ iqooM^
— ^t NORM
F 1


-
'
•—
v^ •
*~ 1
IOM
s
s
s
s
—

•—
•—

r5
i
M
OOK
***, 9
OK
K
-w* — i
""^x
BAL^,
\
IOOK
I
fS»-


r*—
)J
OUT
X

1
+15 -15
r -s
IO-«
io-7
io-«
lO-io








x io-'°
IX 10-'
x io-7
i •». io-«
_5
x IQ-*
IK
9.09K
IK

n,
T

.
                                                                   .0033MF
49.9 K
	
IMF IOMF

49.9K
f 1 J
-H IMF IOMF

L 1 |
IOOMF| {
1-, '

LI I ,
LJ 1
5IIK
5IIK ~
0033H,-5"
MF 1 TR

HORT

                                                                    IK
                                            I SEC
                                                    MO SEC
CALIBRATE
                  FIGURE 10. AMPLIFIER/DEMODULATOR SCHEMATIC

-------
           FIGURE 11
SPECTRAL DISTRIBUTION OF $  EMISSION
Slits:  3.5 - 3.5/2mm col. slit
Concentration :     0.5 ppm

-------
                 FIGURE 12

     SPECTRAL DISTRIBUTION OF S EMISSION
1    Slits: 5-5/ 2mm col. slit
     Concentration :    0.5 ppm

-------
               FIGURE 13

COMPARISON OF MULTI-SLITS AND SINGLE SLITS
Slits: A 3.5 - mulH
     B  3.5-3.5

-------
                          FIGURE 14
    SO  MONITORING WITH MULTI-SLITS IN DIFFERENTIAL MODE
BAIRD-ATDMIC
    Slits:  3.5/multi-slits
    Concentration:  lOppb
f.

-------
Figure  15.  Modified SO- Analyzer (Right Side)

-------
Figure 16.  Modified SCL Analyzer (Left Side)

-------
Figure 17.  Modified SCL Analyzer (Top View)

-------