STANDARD OPERATING PROCEDURE:

 GC/MS ANALYSIS OF HSH, SEDIMENT AND WATER
       FOR VOLATILE ORGANIC COMPOUNDS
       Tammy Jones and Michael  Hiatt
               USEPA Region 9
              August 19, 1983
ENVIRONMENTAL MONITORING SYSTEMS LABORATORY
         LAS VEGAS, NEVADA  89114

-------
                                 Introduction

     The purpose of this standard operating procedure is  to  provide operator(s)
of the Finnigan OWA (Organics in Water Analyzer)  Model  1020  Automated GC/MS,
with a detailed plan for operation, maintenance,  quality  control  and collection
of verified data.  Before this SOP is used, it is assumed the operator  has  read
Finnigan manuals 10000-90080 (OWA Operators Manual,  June  1979)  and 10000-90140
Rev A (OWA Service Manual, February 1981).

     The operator must also read this document in its entirety before analyzing
unknown samples.  All procedures for standard operation of this instrument  must
be followed; at no time, without prior notification  from  operator's supervisor
      f
or project director, should an operator disregard procedure  given, or referred
to, in this document.
                                       n

-------
                                    CONTENTS


                                    "Sections
   I.  Tuning GC/MS (OWA) for Volatile Analysis
  II.  Points for Documentation
 III.  Standard Calibration and Quantitation
  IV.  Standard Preparation
   V.  Sample Storage- Water-Fish-Sediment
  VI.  Sample Pre-Preparation for Fish
 VII.  Sample Analysis and Quantitation
VIII.  Tekmar/Autosampler Operation
  IX.  Assembly of Data Packages
   X.  System Software
  XI.  Construction of Reverse Libraries
 XII.  Quality Assurance
XIII.  Maintenance
                                     Tables
   1-1.  Multiple Ion Parameters
   1-2.  GC Parameters
   1-3.  BFB Ion Abundance Criteria
 III-l.  MDL for Volatile and Semi-Volatile Priority Pollutants
 111-2.  Quantitation Errors and Corrective Steps
  XI-1.  Table of Libraries
XIII-1.  Maintenance Checklist
XIII-2.  Electronic Troubleshooting Chart
XIII-3.  Chromatography Troubleshooting Chart
XII1-4.  Troubleshooting Chart for Analyzer Module


                                    Figures


XIII-1.  Mass Spectrometer with Cover Removed
XIII-2.  Removing the Analyzer
XIII-3.  Mass Analyzer                                                    :
XIII-4.  Exploded View of Ion Source
XIII-5.  Alignment Fixture Base
XIII-6.  Ion Source with Filament Assembly Installed
                                                                    (continued)

                                      iii

-------
                              Figures (Continued)
XIII-7.  Ion Source Prior to Installation of Collector
XI11-8.  Ion Source with Collector Installed
XIII-9.  Electron Multiplier
XI11-10. Closeup of Multiplier Housing
XIII-11. Rear View of Vacuum Manifold
XI11-12. Ion Source Circuit Board  "
XIII-13. Alcatel Pump
XII1-14. Turbomolecular Pump
XIII-15. Disk Drive
XI11-16. Air Filter Removal
                                  Appendices
   A  Safe Handling of Compressed Gases in the Laboratory
   B  Emergency Spill Procedures
                                    iv

-------
                                                               Section No. 	I__
                                                               Revision No. 	
                                                               Date:  08/19/83
                                                               Page   1 of  10~

                 I.  TUNING GC/MS (OWA) FOR VOLATILE ANALYSIS

A.   Electrometer Zero:

1.   Enter Program MT  (>MJ_CR*)
2.   Call Scan N (See Table I) (MT  N CR)
3.   Enable Scan N  (MT:  SE CR)
4.   Vary Electrometer Zero value to give average Noise intensity of  3-6  in
     window 122 and in window 452. (MT: 13, value between -6 and  +6)

     Note:  If Electrometer Zero cannot be lowered to these values Preamp
     must be ajusted.  This is described in OWA Service Manual  VI pp. 1-3
       *
5.   Store correct Electrometer Zero value (MT:  Kl^ CR)

B.   Sensitivity:

1.   (> MT CR)
2.   Call Scan P (See Table I for parameter) (MT:  P CR)
3.   Enable Scan P  (MT:  SE CR)
4.   Adjust ion energy value to give an approximate intensity  of  32,000  for
     69 m/e (MT:5, value of 1 to 10)
*CR = Carriage Return.

-------
                                                               Section No.	I__
                                                               Revision No.
                                                               Date:   08/19/8T
                                                               Page   z of   10"
                      TABLE 1-1.  MULTIPLE ION PARAMETERS
 =============================================================================

                                                         Scan List
 Multiple Ion (MI) Scan Parameters -                 N        P        B


Total Scan Time (sec.)                           0.10      0.15     0.16

Threshold                                           111

No. of Mass Intervals                               2         44

Interval 1-
Low Mass
High Mass
Time (sec.)
Interval 2-
Low Mass
Hi,gh Mass
Time (sec.)
Interval 3-
Low Mass
High Mass
Time (sec.)
Interval 4-
Low Mass
High Mass
Time (sec.)
119.50
125.50
0.03

449.50
455.50
0.03
• •



67.50
70.50
0.03

217.50
220.50
0.03

412.50
415.50
0.03

500.50
503.50
0.03
47.50
50.50
0.03

73.80
76.80
0.03

94.50
96.50
0.03

172.50
175.50
0.03
===============================================================================

-------
                                                              Section No. _ I_
                                                              Revision No. _
                                                              Date:  08/19/83"
                                                              Page   3 of  10"
     Note:  If Ion energy cannot be Increased within the 1 to 10 value it may

     be necessary to increase electron multiplier  voltage.  This should be done

     with the awareness of the lead mass  spectrometrist and recorded In the
                              it

     injection log book.  .  .  •>
                              i



5.   Store correct ion energy value (MT:   K0 CR)



C.   BFB Calibration:



     Note:  2ul of pure BFB is added to the FC43 calibration bulb.  This gives

     an approximate intensity for 95 m/e of 6,000  to 10,000 in Scan B or 174

     m/e is < 2/3 176 m/e.  Then 95 m/e <2000  an. additional 2ul BFB should be

     added to the FC43.




1.   Enter Program MT  (>MT CR)

2.   Call Scan B (See Table I for parameters)   (MT:  SB CR)

3.   Enable Scan B  (MT:  SE CR)

4.   Adjust scan numbers to 5-way  (MT: J5 CR)

5.   Adjust lense voltage to maximize intensity of 219 m/e on scope (MT:  8, 1

     to 15)



6.   Adjust ion program to make 174 m/e 50 to  70%  of 95 m/e.  (MT:  7, 0 to

     20)

-------
                                                             Section  No. 	I_
                                                             Revision No. 	
                                                             Date:  08/19/83
                                                             Page  4 of  10~
7.   Adjust ion energy to make 75 m/e 40 to 60% of 95 m/e (MT:   5, 1  to 10)

     Note:  Ion energy should not be increased appreciably as the sensitivity
     adjustments maximized ion energy value in Section I.B.

     Resolution should be adjust (pots only or rod drive board)

     res. cal. pot       adjusts resolution for 174/175 m/e
     MI mass. res. pot   adjusts resolution for 219/220 m/e
     TP14 GND (A) pot    adjusts resolution for 69/70 m/e

8.   Adjust resolution of 219/220 m/e to give a 30-50% valley.
9.   Adjust resolution of 174/175 m/e to give a 5-10% valley.
10.  Adjust resolution of 69/70 to over resolve.
11.  Repeat steps 7-9 until all resolutions are within their respective limits.
12.  Repeat steps 4-6 until all values are within their respective limits.

     Note:  Repeat steps 7-9 and 4-6 until all criteria are met.

13.  Store correct ion energy  ion programing and lense voltage  (MT:   KO/ CR)
14.  Hardcopy present values   (MT:  to)
15.  Exit program  (MT: £2.

-------
                                                             Section  No. 	l__
                                                             Revision No.
                                                             Date:  08/19/83
                                                             Page   5 of   10"
0.   Mass Calibration:

1.   Enter calibration program (> CAH;>.)

     Note:  This will automatically mass-calibrate system and hardcopy calibra-
     tion limits.  After calibration program is complete computer will  request
     scan parameter data.  Enter previous scan list parameter used for BFB
     performance evaluation as recorded in injection log book.

2.   Hardcopy FC43 spectrum  (>LH FC43#1;LHI2 CR)

E.   Acquisition rate setting.  Normally the following procedure will  adequately
     adjust the aquisition rate such that BFB and sample analysis can  be
     accomplished with the same threshold and minimum area settings.   Adjust-
     ments to this procedure may be necessary in the instance of new or decaying
     eletron multiplier.

F.   BFB Performance Confirmation:

     BFB is injected and chromatographed using GC-3 (See Table 1-2).   Final
     relative abundance and isotope resolution is to fulfill  criteria  in Table
     III, list B.

1.   Set injector system to Auto

-------
                                                             Section  No. 	I_
                                                             Revision No. 	
                                                             Date:  08/19/35"
                                                             Page   6 of  iff"
2.   Call GC parameters 3 (> GC; 3 CR x 12)




3.   Set scan rate to 3 sec.
                              t.

                              •i

4.   Scan parameters threshold land minimum area are adjusted In Section  1C  so
                              .1
     that BFB performance file should be < 100 sectors.   This is adjusted on  a

     "day after" basis if BFB criteria are fulfilled (i.e.,  if BFB  criteria is

     fulfilled and file sector usage is >100, the minimum area will  be

     increased for the next BFB file aquisition.  This can be checked as

     threshold, minimum area and sector usage are recorded in the injection log

     book. \


       f                       «

     Set up Aquisition file for BFB.




     Note:  Files are named by year, BFB, then the Sequence Number  of BFB

     injection.  Only one BFB file will be recorded for any one 10  hour  day;

     e.g., 82BFB001 first BFB performance file in 1982.

-------
                                                             Section No.	I_
                                                             Revision No. 	
                                                             Date:  08/19/83
                                                             Page   7 of  10~
                           TABLE 1-2.   GC  PARAMETERS
GC Parameter
Zone Temperature
Initial Temperature
Final Temperature
Initial Time
Ramp Rate
Final Time
Separator Oven
Temperature
Manifold Temperature
Current -Injection Mode
Splitless Injection
File 1
225°C
60°C
175°C
3 min.
8°C/min.
24 min.
225°C
80°C
<'P
___
File 3
225°C
70°C
165°C
3 min.
30°C/min.
10 min.
225 °C
80°C
C
40 sec.
==============
File 4
225°C
-90 °C
60°C
0 min.
30°C/min.
10 min.
225°C
80 °C
4
60 sec.
:================
File 5
225°C
160°C
200°C
0 min.
6°C/min.
10 min.
225°C
80°C
C
0 sec.
 Time                               '

Turn Fil/Mult Off
 for	sec.  After
 Injection               0 sec.      200 sec.         90 sec.          0  sec.

Current Fil/Mult Mode      MM              M              M

-------
                                                             Section No.	I_
                                                             Revision No. 	
                                                             Date:  08/19/ST"
                                                             Page	8 of  10
5.   Type Acquisition of file (>AC XBFBY#60)

     where x  =  last two digits of year and
           Y  =  Sequence Number for BFB

6.   Type amount BFB Injected (sample:   20ng  BFB  CR)

7.   Type initials of operator (> operator:   >J

8.   Hit return key until file information prompts  are completed.

9.   When screen indicates GC is ready  inject 20ng  BFB -  (See Preparation  of
     Standards) and press GC start button.

10.  At completion of file acquisition  compare BFB  spectra  with  that  of  Table
     1-3, column B.

11..  Hardcopy spectra of BFB using (Ltfspectra No.)  single and/or average
     spectra (LAH CR)

12.  Ensure that injection log book is  completed  for  file entry.

13.  Store correct ion energy ion programing  and  lense voltage
     (MT:  Kl CR)

-------
                                   TABLE 1-3.  BFB ION ABUNDANCE CRITERIA
23:133333333333333333333 3333333333333 = 3333:53333.33.2 3 33 333s = 33333333333333333S33BS33333333333a3:s333;3S3338
                                        Ion Abundance Criteria
 Mass
B
                                           15 to 40% of base peak

                                           30 to 60% of the base peak

                                           Base peak, 100% relative
                                           abundance

                                           5 to 9% of the base peak

                                           Less than 2% of mass 174
 50       20 to 40 pet of mass 95

 75       50 to 70 pet of mass 95

 95       Base peak, 100 pet relative
          abundance

 96       5 to 9 pet of mass 95

173       Less than 1 pet of mass 95


174       70 to 90 pet of mass 95


175       5 to 9 pet of mass 95

176       70 to 90 pet of mass 95


177       5 to 9 pet of mass 95            5 to 9% of mass 176              5 to, 9% of mass 176
32133S23333333333333383333333333=33333333333338333333333SS3S3333833333S333333333333«3S33333333333«»3Z
A Fed. Regist. 1979, 44 (No.233), 69532-69539.
B Olynyk, P., W. L. Budde, and J. U. Eichelberger, 1981.  Simultaneous Qualitative and Quantitative
  Analyses.  I.  Precision Study of Compounds Amenable to the Inert Gas-Purge-and-Trap Method.
  Journal of Chromatographic Science.  Vol. 19.  July, 1981. /IFB's WA 83-A093 and WA 83-A094
C Spies, David N.  "Determination of Purgeables Organics in Sediment Using a Modified Purge and
  Trap Technique"; U.S. Environmental Protection Agency Region  II:  Edison, NJ, October 10, 1980.
                                           Greater than 50% of the base
                                           peak

                                           5 to 9% of mass 174
                     15 to 40% of the base peak

                     30 to 60% of the base peak

                     Base peak, 100% relative
                     abundance

                     5 to 9% of the base peak

                     less than 1%- of the base
                     peak

                     greater than 50% of the
                     base peak

                     5 to 9% of mass 174
                                           Greater than 95%, but less than  greater than 50% of the
                                           101% of mass 174                 base peak
                                                                                                        •O O 30 «/>
                                                                                                        ot ot (9 n
                                                                                                        «o r* < o
                                                                                                        <£>
                                                   o§3
                                                   oo   z
                                                    . z o
                                                   H- O "

-------
                                                             Section  No. 	I_
                                                             Revision No.
                                                             Date:  08/19/85"
                                                             Page   10 of   10
14.  Hardcopy present values  (MT:   H CR)



15.  Exit program  (MT:  E CR)

-------
                                                             Section No.  II
                                                             Revision No.  0
                                                             Dat^:  08/19/83"
                                                             Page  1  of   4~
                         II.  POINTS FOR DOCUMENTATION
                           (or Writing the Daily Log)
A.   Introduction:

1.   Each day a new page(s) in a sequentialy numbered log  book  (the Daily Log)
     is to be used for the recording of the day's  events.

2.   The page is to be titled with the appropriate data  such  as title  book
     number, project and date.

3.   The first entry is to be numbered one, with every new entry therafter
     labeled sequentially.  After the number of entry, the time of entry, and
     then the initials of the entry maker are to be written,  ex:   1) 8:10.

4.   Next the entry is then to be written.

B.   Writing the Entry:

1.   When a different procedure is done a statement(s) defining the objective
     should be included.
          i
          i
          /
     a.  /What is the objective.

-------
                                                             Section No.   II
                                                             Revision No.   0
                                                             Date:   06/01/SF
                                                             Page  2  of    4~
     b.   How Is the objective to be obtained.

     c.   What we do expected results.

2.   The performance standard should be noted In the dally log.
     ex:  Time obtained and BFB Federal Register criteria were all  met.

3.   Standard preparation will be documented In Dally Log by SOP.

     ex:  1.   Internal standard prepared by SOP XX.

        '  2.   VOA Standard prepared by SOP YY.

          3.   (or)   Standards prepared by SOP XX;YY.

4.   Standard runs will be described (by concentration) In the dally log and
     referenced to the Injection log.  A chart should be made to establish the
     cone, to the sample, also included in the chart should be the time of
     acquisition, integration and response list.

-------
                                                             Section No.   II
                                                             Revision No.   0
                                                             Date:   06/qi/^T"
                                                             Page  3  of    4~
                         Table of Daily Runs (Example)
                                  'Standards

         Sample Number     ABC   A A         Acq.      Int.     Resp.  List*

          XXVOA000         150   1000       10:00    10:20       10:45
                                                                (create)

     *It should be noted whether or not the stdls  were created or appended to
      that day response list or the previous day's.  Or if the std.  run was
      not acceptable for either creation or appendation why it was not.  An
      example could be due to the response factor and why this point wasn't
      acceptable for the list.
5.   The preparation of the blank water should also be noted in the log - not

     daily but whenever new water is boiled to be used.
      *



6.   Reagent blanks, samples, duplicates   spikes of samples will  all  be noted

     in the daily log and referenced to the injection log.   They can be added

     to the table of daily runs, as noted above, with the standard runs.



7.   A statement about the quantitation method will be entered after the

     response lists are completed.

                         \


          Example:  A three point linear quantitation report will  be used to

                    report todays analysis.



8.   Any abnormalities concerned with the sample analysis will be noted in the

-------
                                                             Section No.   II
                                                             Revision No.~"TT
                                                             Date:   06/01 /{%""
                                                             Page  4  of    f~
     injection by and the daily log.  One entry can be referenced to the other

     log.



9.   Any instrument abnormalities should also be noted.



10.  After an investigation has been done, a statement(s) should be Included to

     summarize findings; also, perhaps a speculatory statement could be

     included if unexpected results are found.



C.   Finishing the Log Book



1.   Portions of the daily log not used will be filled with X and signed.



          Example	
                               XXXXXXXXXXXXXXXX
                                  Signature
2.   Each page should be signed and dated at the bottom of the page each day.

-------
                                                            Section No. Ill
                                                            Revision No.  0
                                                            Date:  08/19/35"
                                                            Page   1 of  12"
              IV.  STANDARD COMPOUND CALIBRATION  AND QUANTITATION
                             • i
A.  Prepare calibration standards at a minimum of three  concentration  levels
    for. each parameter.  One of the calibration standards  should be at a
    concentration near, but above, the MDL (See Table  1) and  the other concen-
    trations should correspond  to the expected range of  concentrations found  in
    real  samples or should define the working range of the GC/MS system.   IFB
    requires 40 ng/ml , 20 ng/ml, and 4 ng/ml  calibration curve.

B.  Quantisation of Standards

If the file is a standard run the individual  libraries are quantitated:

          WQ_ File name, Lib entry #'s, unknown (CR)

After the Quantitation list has been written check to  see  that all the compounds
have been found.  If some are missing, reset the  corresponding library entry  to
reflect updated relative retention time and retention  time.  Then  enter  IDOS
(>Diag
Rename Quantitation list:
     ex:  (+ Re 8100A542.Q1, \U81VOA642.QL CR)

-------
Section No. Ill
Revision No. ~TT
Date:  08/19/83"

TABLE III-l. MINIMUM C
SEMI-VOLATIi
Parameter
- (yg/D
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.
Acenaphthene
Acrolein
Acrylonitrile
Benzene
Benzidine
Carbon tetrachloride
Chlorobenzene
1,2,4-trichlorobenzene
Hexachlorobenzene
1,2-Dichloroethane
1 ,1 ,1-trichloroethane
Hexachloroethane
1,1 -di chloroethane
1,1,2-trichloroethane
1,1,2,2-tetrachloroethane
chloroethane
bis (2-chloroethyl )ether
2-chloroethyl vinyl ether
2-chloronaphthalene
Page 	 Z of 12
JETECTABLE LIMITS FOR VOLATILE AND
.E PRIORITY POLLUTANTS
======= ======================================
IFB Our
Value Value
10
100
100
5
40
5
5
10
10
1
5
10
5
5
10
10
10
10
18
3
20
11
1
30
2
1
3 '
3
1
1
3
1
2
1
1
2
4
3
       (continued)

-------
Section No. Ill
Revision No.  0
Date:  08/19/8?"

TABLE III-l. Continued.
Parameter
. (ug/D
20. 2,4,6-trichlorophenol
21. 4-chloro-3-methy1 phenol
(para-chloro-meta-cresol )
22. chloroform
23. 2-chlorophenol
24. 1,2-dichlorobenzene
25. 1,3-dichlorobenzene
26. 1,4-dichlorobenzene
27. 3,3'.-dichlorobenzidine
28. 1,1-dichloroethane
29. trans-l,2-dichloroethene
30. 2,4-dichlorophenol
31. 1,2-dichloropropane
32. trans-l,3-dichloropropene
33. cis-l,3-dichloropropene
34. 2,4-dimethlphenol
35. 2,4-dinitrotoluene
36. 2,6-dinitrotoluene
37. 1,2-diphenylhydrazine
38. Ethyl benzene
39. Fluoranthene


IFB
Value
10
10
5
10
10
10
10
20
5
5
10
10
5
5
10
20
20
20
5
10
Page 3 of 12

5SSSSSSSSSSSSSSSSSS
Our
Value
9
5
1
3
6
8
7
5
1
2
6
1
3
4
4
9
7
NA
1
2
       (continued)

-------
Section No. Ill
Revision No.  0
Date:  08/19/85"

TABLE I I 1-1. Continued.
========================================
Parameter
.(ug/D
40. 4-chlorophenyl phenyl
ether •
41. 4-bromophenyl phenyl
ether
42. bis(2-chloroisopropyl)ether
43. bis(2-chloroethoxy)-methane
44. methylene chloride
45. chloromethane
46. bromome thane
47. brdmoform
48. bromodi chloromethane
49. chlorodibromomethane
50. hexachlorobutadiene
51. hexachlorocyclopentadiene
52. isophorone
53. naphthalene
54. nitrobenzene
55. 2-nitrophenol
56. 4-nitrophenol
57. 2,4-dinitrophenol
58. 4, 6-dinitro-2-methyl phenol
59. N-Nitrosodimethylamine


==================
IFB
Value
10
10
20
20
5
10
10
10
5
5
10
10
10
10
10
20
50
50
20
NA
Page 	 4 of 12

=====================
Our
Value
1
1
7
7
2
2
1
2
1
2
7
4
6
2
7
6
9
10
9
10
        (continued)

-------
Section No. Ill
Revision No.  0
Date:  C3/19/8T"
TABLE III-l. Continued.
====================================:
Parameter
(yg/D
60.
61.
62.
63.
64.
65.
66.
67.
68.
69.
70.
71.
72.
73.
74.
75.
76.
77.
78.
79.
80.
N-nitrosodiphenylanrine
N-nl trosodi propyl ami ne
Pentachlorophenol
Phenol
81 s(2-ethylhexyl) phthalate
Benzyl butyl phthalate
Di-n-butyl phthalate
Di-n-octyl phthalate
Oi ethyl phthalate
Dimethyl phthalate
Benzo(a)anthracene
Benzo(a)pyrene
Benzo(b)fluoranthene
Benzo(k)fluoranthene
Chrysene
Acenaphthylene
Anthracene
Benzo(gh1)perylene
Fluorene
Phenanthrene
Dibenzo(ah)anthracene
IFB
Value
10
10
10
10
10
10
10
10
10
10
10
20
20
20
20
10
10
20
10
10
20
Page 5 of 12
:=s========c==r ===============
Our
Value
7
9
9
4
2
2
2
3
3
2
2
2
6
(6)
2
2
2
5
4
3
6
       (continued)

-------
Section No. Ill
Revision No.  0
Date:  08/19/8?"
TABLE III-l. Continued.
==================== ====s===================z=zss==ss
Parameter IFB
. (ug/D . Value
81.
82.
83.
84.
85.
86.
87.
88.
89.
90.
91.
92.
93.
94.
95.
96.
97.
98.
99.
100.
101.
Indene(l,2,3-cd)pyrene
Pyrene
Tetrachloroethene
Toluene
Trichloroethene
Vinyl chloride
Oieldrln
Chlordane
4 ,'4 '-DDT
4, 4 '-DDE
4,4'-DDD
Endosulfan I
Endosulfan II
Endosulfan sulfate
Endrin
Endrln aldehyde
Heptachlor
Heptachlor epoxide
a-BHG (llndane)
b-BHC
d-BHC
20
10
5
5
5
10
0.005
0.050
0.010
0.005
0.010
0.005
O.OOg
0.010
0.005
0.010
0.005
0.005
0.005
0.005
0.005
Page 	 6 of 12
============C==M=CS=C=3=;
Our
Value
NO
2
1
1
2
1
2

1
2
2
3
8

8

3
1
3
4
1
       (continued)

-------
Section No. Ill
Revision No.  (L
Date:  08/19/55"
Page 7 of 12
TABLE III-l. Continued.
======================================================== ====ss:======s=cs===s=zs
Parameter IFB Our
(yg/1) Value Value
102.
103.
104.
105.
106.
107.
108.
109.
110.
111.
112.
113.
114.
115.
116.
117.
118.
119.
120.
121.
122.
====
g-BHC (lindane)
Acetone
Aniline
Benzole Acid
Benzyl alcohol
2-Butanone
Carbondisulfide
4-Chloroanillne
Dibenzofuran
2-Hexanone
2-Methyl naphthalene
4-methyl -2-pentanone
2-methyl phenol
4-methyl phenol
2-nitroanillne
3-nitroanillne
4-nltroanillne
Styrene
2,4,5-trlchlorophenol
Vinyl acetate
o-Xylene
0.005
5
5
100
20
5
1
50
10
5
20
5
5
5
100
100
100
5
100
5
5
3
8
4
6
2
4
2
4
2
2
3
2
9
2
2
8
4
4
6
NA
2
============================

-------
                                                             Section  No.  Ill
                                                             Revision No.  0
                                                             Date:  08/19/35"
                                                             Page   8 of   12
           TABLE II1-2.
                    QUANTITATION ERRORS AND CORRECTIVE STEPS FOR
                    COMPOUNDS KNOWN TO BE PRESENT
===========================================================scco==c===e=========
      Error
                        Observation
                                       Correction
 1.
 2.
Wrong peak
integrated.
No peak
integrated,
1.
Multiple "hits" in
reverse library status
report.
    Compound not in reverse
    library window.
                       Compound in "window" and
                       peak shape is normal (no
                       attempt to integrate).
                       Compound in "window",
                       peak shape normal,
                       (attempt is made to
                       integrate).

                       Compound in "window"
                       peak shape not normal,
1.  Narrow window width.
2.  Raise fit threshold.
3.  Make-reverse library mass
    spectra more resistive
    (i.e. add masses).1

1.  Widen window width.
2.  Change r.r.t to place
    compound in window center.

1.  Lower fit threshold.
2.  View compound reverse
    library mass spectra (VI
    command) and identify any
    "obscured mass-ion peak.
    Remove obscured mass
    fragment.1

1.  Narrow window width.
                                             1.  Compound must be inte-
                                                 grated using WQ file,
                                                 reverse Lib.OWAIN1.

                                             2.  Compound can be manually
                                                 quantitated by QH of mass
                                                 peak and RP program (all
                                                 steps to be hardcopied).
1 These steps to be done with assist of lead mass spectrometrist.

-------
                                                             Section No. Ill
                                                             Revision No.  0
                                                             Date:  08/19/85"
                                                             Page   9 of  12
Return to executive (+ M CR ):

Requantitate the standard using the library with  the  updated entry(les).

If a standard compound can not  be found in the  reverse  library search with the
corrections, then the error Is  identified and it  will be noted In  the comment
column of the injection log (See Table III-2 for  quantitation errors).

C.  Reverse library quantitation lists are merged by  adding V2. to  Vl_ then V^
     to \a, V4 to Vl_ and finally V5 to VI:

     (? EQL V282VOA642,V182VOA642 CR)
        EQL:  A;E CR
      > EQL V382VOA642.V182VOA642 CR
        EQL:  A;E CR
      > EQL V482VOA642.V182VOA642 CR
        EQL:  A;E CR
      > EQL V582VOA642.V182VOA642 CR
        EQL:A;E CR

0.  Rename Quantitation list to file name:

                                         > Oiag CR
                                         + Co V182VOA642.QL.81VOA642.QL
                                         + M CR

-------
                                                             Section No.  Ill
                                                             Revision No.0~
                                                             Date:   08/19/35"
                                                             Page  10 of   12"
E.   Creating and Updating Response Lists/Calibration Curve

'Adjust library quantisation amount to reflect the amount of compound added,
make sure the correct amounts of the compounds and ISTOS are updated:
     ex:  > MQ 82VOA642 CR
     MQ:  :D;; (amount added), Quantisation report entry number(S);0;;

     Note:  Consecutive compounds of the same concentration can be strung
     together as described by:

          MQ: ?.
       f                   -                    •
F.  Update response  lists:  (MQ:;0;;;R;;;S;;;T;;;>or<(append or create,
dependent upon calibration checks);;E

G.  After the 3-point (or more) calibration curve has been created and updated
then hardcopy updated Quantitation list;

          >MQ 82VOA642 CR
          MQ: H CR

H.  Calibration Curve Check

Check of the calibration curve should be performed once each 8 Jiour shift.

-------
                                                             Section  No.  Ill
                                                             Revision No.5~
                                                             Date:  08/19/8T"
                                                             Page  ll of   12"
These criteria are described in detail  In the instructions for IFB Form £,
Calibration Check.  The minimum response factor for the system performance
check compounds must be 0.300.   If this criteria Is met, the response factor of
all compounds is calculated.  A percent difference of the daily response factor
(each 8 hr. shift) compared to the average response factor from the Initial
curve is calculated.  The average response factor (RF~) should be calculated  for
all compounds.  A system performance check must be made before this calibration
curve is used.  Six compounds (the system performance check compounds) are
checked for a minimum average response factor of 0.300.  These compounds (the
SPCC) are chloromethane, 1,1-dichloroethane, 2-chloroethylvinyl  ether, bromo-
form, 1,1,2,2-tertrachloroethane, and chlorobenzene.  The maximum percent
difference allowed for each compound flagged as  'CCC' in IFB Form X is 20%.
Only after both these criteria are met can sample analysis begin.

I.  Internal standard responses and retention times in all samples must be
evaluated immediately after or during data acquisition.   If the retention time
for any internal standard changes by more than 30 seconds from the latest daily
(8 hr. shift) calibration standard, the chromatographic system must be
inspected for malfunctions and corrections made as required.  Ifrthe extracted
ion current profile (EICP) area for any internal standard changes by more than
a factor of two (-50% to +100%), the mass spectrometric system mus be inspected
for malfunction and corrections made as appropriate.  When corrections are.made,
re-analysis of samples analyzed while the system was malfuncioning is necessary.

-------
                                                             Section No. Ill
                                                             Revision No.  0
                                                             Date:  08/19/85"
                                                             Page  12 of  12
J.  Retention time and EICP area records shall be maintained in the form of

control charts (Exhibit E. Figure 3, IFB WA-83-A094) by the laboratory as a

part of its internal quality control.  Internal standard response data shall

be reported on the I.S. Response Verification Data Sheet (see Exhibit B, IFB

WA-83-A094).                                              .

-------
                                                             Section No.  IV
                                                             Revision No.  0
                                                             Date:  08/19/85"
                                                             Page   l of   7"
                        IV.  STANDARD PREPARATION
                              t
A.  The standard stock solution has two aliquots:
                              I
                              , i
     The first 1s made from standard mixtures sold by Supelco (Supelco,  Inc.,
Bellefonte, Pennsylvania) under the name Purge A and Purge B.  Each  of these
comes in 1 ml vials with a concentration of 0.2 mg/ml for each of the compounds
present in Purge A and Purge B.   In addition 2-cholorethyl  vinyl  ether and
trichlorofluoromethane is added to the first aliquot.  The final  concentration
of the first aliquot is 50 ng/ul except for 2-chloro ethylvinyl  ether which  is
100 ng/,wl.
                                   -f
     The second aliquot contains four compounds.  These are obtained in
separate ampules.

     Methyl Bromide      (Concentration 1510 pg/ml)
     Vinyl Chloride      (Concentration 7300 yg/ml)
     Chloromethane       (Concentration 5350 pg/ml)
     Chloroethane        (Concentration 1080
These compounds are stored in lab freezer, Rm. 49, and the contents of each
ampule are transferred to a separate gas-tight syringes.  Measured amounts of
each are added to make a 1 ml solution with a final concentration of 50 ng/yl.

-------
                                                             Section  No.   IV
                                                             Revision No.  0
                                                             Date:  08/19/85"
                                                             Page   2 of   T
The solvents used in the preparation of Purge A & B,  Purge C are methanol  and
ethylene glycol.  In the preparation of Acrolein and  Acrylonytrile,  and
internal standards for daily use, b.oiled-distilled-purged H20 is the solvent.

Procedure for making daily solution of Purge C* for vacuum distallation  and
purge and trap.

1)   Pipette 904.4 yl of ethylene glycol to gas-tight syringe used for the
     daily preparation of Purge C.

2)   Add 33.1 yl of methyl bromide (concentration 1510 yg/ml).
      f
3)   Add 46.3 yl of chloroethane (concentration  1080 yg/ml).

4)   Add 9.3 yl of chloromethane (concentration 5350  yg/ml).

5)   Add 6.9 yl of vinyl chloride (concentration 7300 yg/ml).

Internal Standards for vacuum distillation and purge  and trap.

A)   The internal standard solution is made of five compounds.  Three of them
     are sold by Supelco (Supelco, Inc., Bellefonte,  Pennsylvania) in 1  ml.
     aliquots with a concentration of 20 mg/ml.  The three compounds present
'Calculations for the amounts added are found in the standards log book
 (1982) for Region 9, Pages 1-10.

-------
                                                             Section No.  IV
                                                             Revision No.  0
                                                             Date:  08/19/IT
                                                             Page   3 of   7"
     are the following:

          Bromochloromethane
          1,4-Dichlorobutane
          l,2-Chloro-2-bromopropane

     The two remaining internal standards are obtained in the pure  form,  they
are:

          Ds-Benzene        density 0.851 g/ml*
          Ds-Toluene        density 0.945 g/ml*

     A 1 ml stock solution with a concentration of 2 mg/ml is made  in  the
following manner:

1)   Pipette 895.6 pi of methanol into a clean 1 ml reaction vessel.

2)   Pipette 100 pi of Supelco internal standard solution into the  same
     vessel, or

3)   Add 2.1 yl D5-Benzene with a 10 pi syringe into vessel.

4)   Add 2.1 ul Ds-Toluene with a 10 pi syringe into vessel.
*Refer to density measurement in the 1982 Standards Log Book for Region 9,
 page 2.

-------
                                                            Section No.  IV
                                                            Revision No.  0
                                                            Date:  08/19/35"
                                                            Page   4 of   7~
B)   The stock internal standard solution Is used  to  prepare an 80 ng/5 yl
     aliquot for use in sample analysis.   The following procedure 1s used:
                             tt
     1)   Pipette 992 yl of boiled-distilled-purged t^O into a 1 ml crimp
          top vial.           ,

     2)   Add 8 pi of the stock internal  standard  solution  (2 mg/ml) to the
          crimp top vial.

     3)   Crimp a top to the vial, forming a good  seal.

Ethylene Glycol

1)   Pour approximately 50 ml of pure ethylene glycol into  the sample  holding
     bulb on the volatile organic extraction rig.

2)   Extraction of any remaining contaminants is carried  out by  following:
     the first 20 minutes of the volatile organics extraction procedure  (see
     Section III).

4)   The sample bulb is then removed and capped using a Buna-n-ring  and  glass
     stopper.  The bulb is labeled, dated, and kept at room temperature  for
     future use in preparation of standards.

-------
                                                             Section No.  IV
                                                             Revision No.~"TT
                                                             Date:  08/19/SF
                                                             Page   5 of   T~
     Pure P-Bromofluorobenzene is diluted with methanol  (spectro grade) to

make a solution with a final concentration of 40 ng/yl.



Procedure:*



     1)   Take 50 ul of concentrated/pure BFB density and dilute to 5 ml with

          methanol.



     2)   Add 13.4 yl of solution made in step 1 to a clean 5 ml volumetric

          flask.



     3)   Add methanol to solution in step 2 to fill volumetric to the  5 ml
       *
          mark.



Boiled-Distilled-Purged H20



     1)   Distilled tap water is re-distilled (Corning mega-pure distiller).

          Re-distilled water is collected in a 1-liter erlermeyer flask.



     2)   Water is then boiled for 1/2 to 1 hour on a hot plate.



     3)   The re-distilled boiled water is transferred into a 1-liter round-

          bottomed flask that has an o-ring neck adapter.
^Calculations  for the amounts added are to be found in the 1982 Region 9
 Standards Log Book pages 1-10.

-------
                                                             Section  No.   IV
                                                             Revision No.  0
                                                             Date:  08/19/33T
                                                             Page   6 of   7
     4)   The round-bottom flask is connected to the volatile  extraction  rig.



     5)   Extraction of the remaining contaminants Is carried  out  by following

          the first 20 minutes of the volatile organlcs  extraction procedure

          (see Section III).



     6}   The water Is removed and capped with a Buna-n-ring.   A source clamp

          Is used to tighten the seal, then It Is labeled,  dated and kept at

          room temperature for daily use.



Preparation of Purge A and B; Daily STD Soln. for purge  and trap analysis


       *

Method A:



     1)   Pipette 154 yl MeOH into a std. vial (labeled  STD soln.  A and B).



     2)   Pipette 100 yl Purge A (Supelco) into same vial.



     3)   Pipette 100 yl Purge B (Supelco) into same vial.



     4)   Pipette 40 yl Purge H (Radian) into same vial.



     5)   Place lock type cap (green in-open or red in-closed).

-------
                                                             Section No.   IV
                                                             Revision No.   0
                                                             Date:   08/19/8F
                                                             Page   7 of    7'
     6)   4 yl 2-chloroethylvinyl ether (10,000 ng/yl) into small  vial, using a

          syringe.

     7)   2 yl-Trichlorofluoromethene (10,000 ng/yl) into small  vial, using a

          syringe.



Method B (use when supply of Purge A and B are low)



     1)   40 yl MeOH



     2)   25 yl Purge A



     31   25 yl Purge B



     4)   lOyl Purge H



     5)   Cap vial
                               *


     6)   1 yl 2-chloroethylvinyl ether (10,000 ng/yl)



     7)   0.5 yl Trichlrorfluoromethene (10,000 ng/yl)

-------
                                                             Section No.   V
                                                             Revision No.  0
                                                             Date:  08/19/83
                                                             Page   1 of   1~
                   V.   SAMPLE STORAGE - WATER - FISH - SEDIMENT

•A.    Duplicate water  samples are received In 40 ml septum vial containers and
      are  stored  in a  refrigerator placed in a volatile organic-free room.
      Samples  are to be analyzed within 14 days of_cojj£ctloju.

B.    Upon receipt  of  a solid fish sample a 10 g aliquot is added to a 100 ml
      sample chamber.

      1.    The aliquot should represent homogeneous materials and omitting
           rocks  and obvious organic material.

      2.    The sample  chamber is sealed using a Buna-N o-ring, o-ring cap, and
           pinch  clamp.  Care is taken to ensure an air-tight seal.

      3.    The sample  chamber and sample is stored in a freezer (»-17C) placed
           in  a volatile organic-free room.

C.   Sediment  samples  are received  in sealed mason jars and are stored in
volatile  free walk in refridgerator.  Samples are jo be_analyzed within 14 days
of  collection.

-------
                                                             Section No.  VI
                                                             Revision No.  0
                                                             Date:  08/19/8T"
                                                             Page   1 of   3"
                      VI.  SAMPLE PRE-PREPARATION  FOR  FISH



A.   Homogenization:



     This is ideally accomplished the day of,  or the day after,  receipt.   If

     the fish tissue amounts to less than 1/4  pound, then an  industrial grade

     stainless steel Hobart meat grinder is  used to homogenize the  sample.



     Manual, homogenization involves chopping the sample  as  finely as possible

     with the meat cleaver on the chopping board.



     To use the Hobart meat grinder, the Biology Bldg. room must be set up

     first.  At the back of the room, .the main power,  hot and cold  water  are

     turned on.  The thermostat is set to maintain the room at 10-16C.  Cover

     the counter tops beneath the Hobart with  garbage  bags.   Place  the Hobart

     in a garbage bag also.  You will need to  stock the  supplies listed;
          paper towels                  ceramic or stainless  spatulae
          lined garbage can             permanent marker
          scrub pads                    tank of liquid nitrogen
          sponges                       Dewar flask
          alcohol/squeeze bottle        100 ml VOA bulb
          deionized water/squeeze       rubber gloves
            bottle                      250-400 ml beaker
          amber grab sample bottles     meat cleaver
          adhesive dot labels           stainless bone saw
          knives                        cloth gloves
          chop board                    lab soap

-------
                                                        Section  No.  VI
                                                        Revision No.  0
                                                        Date:  08/19/53"
                                                        Page   2 of   3~
After thorough soap and water washing,  the grinder,  its  pieces  and  all
knives and cleavers are then thoroughly rinsed with  deionized water and
                                                                     f-
dried with paper towels.  (The Hobart is placed within a large  garbage
bag for the ease of room clean-up).  Turn on the Hobart  by pressing the
large button at the front and pour about 300-500 ml  of liquid nitrogen
into the Hobart feed.  To this slowly add egg-sized  chunks of tissue
which have been chopped or cut using a  stainless cleaver or bonesaw on a
chopping board.  Cloth gloves should be worn under the rubber gloves
because everything is frozen and cold to the touch.

After adding tissue to the nitrogen-filled grinder,  the  plastic bag
opening is pulled down over the top so that the tissue doesn't  contamin-
ate the roonu  After the noise level lowers slightly and the chunk  has
become powdered, more nitrogen and fish chunks are added and the process
is repeated until the whole fish has been ground into a  fine frozen
powder.  At no time is the tissue allowed to thaw.  Weighing out of a 10
g sample of the frozen fish powder is accomplished at this time on  a
calibrated top-loading balance and is clamped in a 100 ml volatile
organic analysis (VOA) bulb as is blank air from the room.  Note that VOA
bulb must have its opening cleaned of sample bits before it can be
effectively sealed with its cap using the Buna-N o-ring  and the clamp.

The remainder of the fish powder is stored in an amber grab bottle  for
the other analyses.  These three containers are placed back in  the  ice

-------
                                                       Section No.  VI
                                                       Revision No.~TT
                                                       Date:  08/19/8F
                                                       Page   3 of   3~
chest or greenhouse freezer  until they are transported to the freezer in

Chemistry Lab room 26 for processing.  They are transported as before in

an ice chest with dry ice.   The  freezer in Room 49 of Chemistry is kept

at -20 to -IOC.

-------
                                                             Section No. VII
                                                             Revision No.  1
                                                             Date:  06/01/32"
                                                             Page   1  of   5"
          VII.  SAMPLE (FISH,  SEDIMENT,  WATER)  ANALYSIS  &  QUANTITATION

     This section contains the various procedures used for analyzing  and
quantltatlng the various environmental  samples.   There is  a 14  day  turn around
time on analysis of water and sediment.

I.  Water Preparation -

1.  40 ml -septum vial containers are removed  from storage and allowed to
    equilibrate to room temperature just before analysis.
        *
2.  Open one of the sample vials and carefully  pour the  sample  into a 5ml
    syringe barrel to just short of overflowing*  Replace  the syringe
    plunger arid compress sample till the barrel's Teflon tip is adjusted  to the
    5 ml volume mark.

3.  Add 5 ul of the ISTD solution and 80 ng of  each surragate to the  sample
    barrel.  This is done by placing the lOul-syringe needle into the larger
    syringe barrel and depressing the plunger of the smaller syringe.

4.  Attach 5ml-syringe to leuver-lock on Autosampler purging trap (See  sec.	
    on Auto Sampler operation and setup)*  Open valve and  depress syringe
    plunger, emptying 5 ml sample into purge trap.

-------
                                                             Section  No. VII
                                                             Revision No.~T~~
                                                             Date:  06/01/8T"
                                                             Page   2   of   5"
5.  Start purging sequence.  When finished Initiate Acquire program 1f

    operating manually or ASRUN If operating on automatic (see section _ J.

                              /,

II.  Sediment Preparation


                              ,f

1.  Remove mason jar containing sediment sample from storage.   Allow to

    equilibrate to room temperature.



2.  Weigh our lOg of sediment into a 100 ml -sample chamber.



3.  Seal sample chamber to vacuum distillation apparatus using a  Buna-o-ring

    and> vacuum pinch clamp.  Seal at porehole marked "sample."
4.  Add 2 ml of blank H20 thru attaching 5 mi-syringe with 2 ml  of t^O  to  hole

    above valve 7.  Open valve 7 allowing the 2 ml  to flow into  chamber.   Shut

    valve.  Remove syringe.



5.  Place sonicator under sample chamber and raise high enough so water level

    comes to shoulder of sample chamber.



6.  Place special fitted purging (glass) trap to porthole Cj, on vacuum

    distillation apparatus.  Immerse in liquid nitrogen.

-------
                                                              Section No. VII
                                                              Revision No.   1
                                                              Date:  06/01/82
                                                              Page  3  of    5~
 7.  Evacuate glass trap by opening valve at  V£ until  purging  trap  Is emptied.
     Shut valve off.

 8.  Start sonication of sample and open valve ¥5  so that the  sample chamber  and
     glass trap are open only to each other.   Allow to equilibrate  for  5 min.

 9..•  After equilibration open valve v"2.   Now  system is open to vacuum.  Continue
     adding liquid nitrogen to dewar at  Cj as needed.   Allow sample to  come over
     into trap for 15 min.

10.  After 15 (min), trapping is completed.  Shut  all  valves,  turn  off  sonicator
     and, lower dewar from purging trap.   Remove purging trap and  attach to
     special marked place on autosampler.

11.  Once purging trap is attached and sealed add  3 ml of blank HgO.  This H20
     is to contain the amounts of internal standards and surrogates as  mentioned
     in  1.3, this section.

12.  Allow ice in trap to melt.  This can be  done  by gently warming using a heat
     gun or allowing sample to melt at own pace.

13.  After sample has melted proceed as  in 1.5.

-------
                                                            Section No. VII
                                                            Revision No.~T~
                                                            Date:  06/01/52"
                                                            Page   4 of   5~
III.  Fish Preparation

1.  Remove 100 ml sample chamber (as prepared In section  >     . part B) from
    freezer and attach to vacuum distillation apparatus as,noted in II.3.

2.  Continue as in II.3 thru 11.13.

IV.  Sample Quantitation

1.  After 3-pt calibration curve has been established  then sample  acquisitions
    can be quantitated.

2.  Type:  WQ_ file name, Reverse library name(s),  unknown (CR).   This will           )
    start the computer quantitating the sample acquisition.

3.  After quantitation is completed, observe the quantitation  report.   If  the
    amount of compound is greater than the reportable  levels,  and  a search of
    the NBS library confirms the presence of the compound(s)  (i.e. the fit and
    purity are high), then the level of contamination  is  reported  to the lead
    spectrometrist.  He/she willthen determine, on the basis  of his/her
    experience, if the sample needs to be re-run for further confirmation  of
    the presence of such compound(s).  If the lead  spectrometrist  decides  the
    amounts present warrant further confirmation the sample is re-run, possibly
    at a greater dilution; the amount of dilution is determined by the lead

-------
                                                            Section No. VII
                                                            Revision No.  1
                                                            Date:  06/01/5T"
                                                            Page   5 of   5~
    spectrometrist - or else, on the basis of his/her  experience.   If the

    amount of pollutant Is sufficiently  high, then  at  the discretion of the

    lead spectrometrist, the values are  reported immediately to his/her
                             /      ~

    superior.  If it is determined that  there is no apparent reason to re-run

    the sample, then the proper calculations are made  to convert to the units
                             ,(
    reported on the STORET system.




4.  See section on "Assembly of Data Packages."

-------
                                                              Section No. VIII
                                                              Revision No.  0
                                                              Date:  08/19/31"
                                                              Page  1  of   4""
                        VIII.   Tekmar/Autosampler Operation

       This section contains the various  procedures that are .used for operating
  the Tekmar and Autosampler.

  I.   Attach to the Tekmar the number of  purge traps  needed for the analyses.
      The traps are to be slipped onto  the sampling needle and Into the "screw"
      nut.  The nut should be gently turned until tight and so that the trap
      won't slip out.  The carrier gas  flow tube should then gently be Inserted
      Into the traps arm.  The traps are  to be cleaned and dried after each day's
         f
      use.

 II.   The Tekmar and autosampler switches are set  Into their proper positions
      depending upon which mode (Auto or  Manual) Is to be used.  Table 1 contains
      the information to be used in setting the switches on the Tekmar and/or
      Autosampler.

III.   The required standard runs are injected into  traps 1 thru 3  (depending upon
      the number of standards to be used).  The next  trap is to contain the
      purged t^O blank (plus internal standards and  surrogates).   The remaining
      traps are to coircain   e sample aliquots plus  internal standards and   .
      surrogates.  Pu» jmy  '*y then be  initiated on the first trap.

-------
                         Section No.  VIII
                         Revision No.  0
                         Date:   08/19/35"
                         Page  2  of    f~
TABLE 1.

-------
                                                             Section  No. VIII
                                                             Revision No.~~TT
                                                             Date:  08/19/3T"
                                                             Page   3 of   4
IV.  Before GC/MS operation.

      A)   The ASSET program Is called to screen.   In asset only files  3,  4,
           and 5 are used.   File 5 Is  for the  first  sample, fl.le 4  for  the
           middle samples and file 3 for the last  sample (a dummy run). See
           Table 2 for parameters.  The computer will  prompt the user.

     ...B)   ADFILE is called to screen, the computer  will prompt the user.
           Placed Into ADFILE are the  files used In  the autosampler analyses.
           After the last real sample number Is entered a dummy file Is added
           to the ADFILE list.
        *
      C)   XFILE is called to screen,  computer will  prompt the user. The  asset
           files used are placed into  XFILE in proper order.  Use AUTOS5 for
           entry number 1, AUTOS4 for all other entries except the  last entry
           (dummy file) where AUTOS3 is used.

      D)   The autosampler routine is ready for initiation.  ASRUN  is typed
           into the terminal, computer will prompt the user.  Page  through the
           questions (unless changes are necessary).  The computer  will take
           over analyses and quantitations.

      E)   If the computer is interrupted ASRUN will be terminated.  In order
           for the Autosampler acquisitions to automatically continue the
           ADFILE and XFILE editors need to be properly reset.

-------
                         Section No. VIII
                         Revision No.  0
                         Date:  08/19/35"
                         Page   4 of   4~
TABLE 2

-------
                                                             Section No.   IX
                                                             Revision No.  0
                                                             Date:   08/19/51"
                                                             Page  1  of   2
                            Assembly of Data Packages

I.   Batch Reports

     For each day's analyses there should be a batch report containing that
day's various parameters, i.e., BFB tune package and sample report package.

     A.   BFB tune package is to contain a copy of manual tune parameters,
          calibration file (CA), copy of scan list of FC43 scan number 1 and
          number 2, and a copy of the scan list of the BFB analysis.

     B.   Sample report package should contain:

          1)   A RIC with identified peaks labeled and internal standard peaks
               labeled.
  *
          2)   An updated quantitation report with a library status report.

          3)   A line that is 25% of the internal standards will be drawn and
               any peak over or to that line will be searched.

               a)   Both enhanced and raw spectra will be hardcopied.

-------
                                                             Section No.   IX
                                                             Revision No.   0
                                                             Date:   08/19/35""
                                                             Page  2  of    2"
               b)   Forward library searches of NBS library will  also be
                    included.

               c)   The program ESFLS should be used as it will  satisfy the
                    required data la and Ib.

     C.   A xerox copy of that day's injection log will be included.

II.  Completion of Report Packages

     A.   A completed data package will contain all the items listed  as A, B,
          and C in Section I, standard analyses reports, blank water  report,
          and a copy of that day's updated response list.

     B.   Final report sheets will accompany each data package.

          1)   This report sheet will reflect the % of surrogates recovered,
               proper report amounts of compounds found (using proper units)
               and other data as prescribed by the updated IFB.

-------
                                                             Section No. JC	
                                                             Revision No.  0
                                                             Date:  08/19/55"
                                                             Page   1 of   l"
                              X.   System Software



     Files from older and newer Finnigan Software  disc  packs have been integrated

into the Region 9 systems software.   The current sotware that is used as of

8/15/83 is titled 2.06.82.  The files extracted from other master discs are

PREVI.PR (subroueine in ASRUN), WQ.PR, XFILE.PR, ASRUN.PR, ASRUN1.PR, ASSET.MS,

ASSET.PR AND KWQ.PR from disc 1.3.80 and 45EXEC.MS are  from the Finnigan 4510

software, 182.

-------
                                                               Section No.   XI
                                                               Revision No.   0
                                                               Date:   08/19/55"
                                                               Page   1 of   11
                     XI.  CONSTRUCTION OF REVERSE LIBRARIES



A.   Choice of Internal Standards.  Of the four following steps one or more

     of them may be used, It Is up to the discretion of the operator.



     1.   Internal standards are nearest compound of Interest.



     2.   Internal standard behaves chromatographically similar to compound of

          interest.



     3.   Internal standard produces best precision data.



     4.   Internal standard quantitation mass is similar to compound of

          interest.



B.   Library Order



     1.   Internal standard is listed first.



     2.   All other compounds are listed in order of elution time.

-------
                                                               Section No.  XI
                                                               Revision No.  0
                                                               Date:  08/19/31"
                                                               Page   z of  11"
C.   Reverse library entries.


     1.   Quantitation mass selection is chosen on a  basis  of "ruggedness"
          (i.e. minimum of quantisation errors, and/or best-experimental
          precision).

     2.   Mass fragment used in reverse searches begins with all  ions  over  5%
          of base peaks.  Ions are deleted as  "mass problems" are identified
          in actual analysis, see Table IV.

        ,  Note:  Reverse libraries are constructed to be what "works"  best  and
          are updated to reflect maximum precision:  i.e.,  libraries are
          improved to minimize false "hits" but maintain maximum
          identification of compounds at their reported detection limits.

-------
                                                                 Section No.  XI
                                                                 Revision No.  0
                                                                 Date:  08/19/83
                                                                 Page   3 of  ll
                        Tables of Libraries:  XI-1.
LIBRARY EDITOR
C. H2. CL. BR
NAME:      BROMOCHLOROMETHANE
F. T.  :        900  SWW   :
R. R. T:      1. 00  MASS  :
              DATA:  LIBRARYU1:
              M. WT.:  128
                 70
                128.
                        QUAN
                        AMNT
MASS INTENSITY
  49  1000
  91/  100
MASS INTENSITY
  93   322
  95   214
LIBRARY EDITOR
Co. D6
NAME:      D6 BENZENE
F. T.  :       -~800  SWW   :
R. R. T:      1. 00  MASS  :
                 MASS INTENSITY
                 128    705
                 130    958

              DATA:  LIBRARYU1:
              M. WT.:   84
                 60
                 84.
                        QUAN
                        AMNT
MASS INTENSITY
  54    89
MASS INTENSITY   MASS INTENSITY
  82   142        83     45
LIBRARY EDITOR
C7. oa
NAME:      D8 TOLUENE
F. T.  :        850  SWW   :
R. R. T:      1. 00  MASS  :
              DATA:  LIBRARYU1
              M. WT.:  100
                 60
                100.
                        QUAN
                        AMNT
MASS INTENSITY
  98  1000
MASS INTENSITY
  99    95
                 MASS INTENSITY
                 100    729
LIBRARY EDITOR                 DATA:  LIBRARYU1:
C3. H6. CL. BR                 M. WT.:  156
NAME:      2-BROMO-1-CHLOROPROPANE
F. T.   :        850  SWW   :       60     QUAN   :
R. R. T:      1. 00  MASS  :       77.    AMNT   :
MASS INTENSITY
  76    37

LIBRARY EDITOR
C4. H8. CL2
NAME:      1.4-DICHLOROBUTANE
F. T.  :        850  SWW   :
R. R. T:      1. 00  MASS  :
MASS INTENSITY   MASS INTENSITY
  77  1000        78     39
              DATA:  LIBRARYU1:
              M. WT.:  126
                 60
                 90.
                        QUAN
                        AMNT
MASS INTENSITY
  49    92
  55  1000
MASS INTENSITY
  62   136
  64    46
                 MASS INTENSITY
                  75     55
                  90    144
                                                BASE M/E:     49
                                                RET. T:    10:49
    A
80.00
           NG
MASS INTENSITY
132    188
                                                 BASE M/E:    84
                                                 RET. T:   10:55
   80.0/0   NG

MASS INTENSITY
 84   1000
                                                 BASE M/E:    98
                                                 RET. T:   12:18
   80.00   NG

MASS INTENSITY
                                                 BASE M/E:    77
                                                 RET. T:   12:45
                                     80.00   NG

                                  MASS INTENSITY
                                   79    369
                                                 BASE M/E:    55
                                                 RET. T:   14:02
       A
   80.00   NG

MASS INTENSITY
 92     45

-------
                                                               Section No.  XI
                                                               Revision No.0~~
                                                               Date:  08/19/8F


Page 4
of 11
Tables of Libraries XI-1. (Continued).
LIBRARY EDITOR
C. H3. CL
NAME: CHLOROMETHANE
F. T. : 850 SWW :
R. R. T: 0. 73 MASS :
MASS INTENSITY MASS INTENSITY
47 102 49 116
48 40 50 1000
LIBRARY EDITOR
C2. H3. CL
NAME: VINYL CHLORIDE
F. T. : 850 SWW :
R. R. T: 0. 76 MASS :
MASS INTENSITY MASS INTENSITY
61 86 62 1000
LIBRARY EDITOR
C. H3. BR
NAME: BROMOCHLOROMETHANE
F. T. : 950 SWW :
R. R. T: 0. 81 MASS :
MASS INTENSITY MASS INTENSITY
79 190 91 92
81 189 93 231
LIBRARY EDITOR
C2. H5. CL
NAME: CHLOROETHANE
F. T. : 950 SWW :
R. R. T: 0. 83 MASS :
DATA: LIBRARYU1:
M. WT.: 50
200 QUAN :
50. AMNT :
MASS INTENSITY
51 35
52 311
DATA: LIBRARYU1:
M. WT.: 62
200 QUAN :
62. AMNT :
MASS INTENSITY
63 41
DATA: LIBRARYU1:
M. WT.: 94
180 QUAN :
94. AMNT :
MASS INTENSITY
94 1000
95 143
DATA: LIBRARYU1:
M. WT.: 64
160 QUAN :
66. AMNT :
6 BASE
RET.
A
100.00 NG
MASS INTENSITY
7 BASE
RET.
A
100.00 NG
MASS INTENSITY
64 313
8 BASE
RET.
A
100.00 NG
MASS INTENSITY
96 941
9 BASE
RET.
A
100.00 NG
M/E: 50
T: 7:42

M/E: 62
T: 8:07

M/E: 94
T: 8:39
'
M/E: 64
T: 8:50
MASS INTENSITY
  49   334
  51   103
MASS INTENSITY
  63    64
  64  1000
MASS INTENSITY
 66    309
MASS INTENSITY

-------
                                                              Section No.  XI
                                                              Revision No,  0
                                                              Date:  08/19/83"
                                                              Page   5 of  11"
                  Tables of Libraries XI-1.  (Continued).
LIBRARY EDITOR                 DATA:  LIBRARYU1:
C. CL3. F                      M. WT.:
NAME:      TRICHLOROFLUOROMETHANE
                                    10
                                BASE M/E:   101
                                RET. T:     9:15
F. T. : 800 SWW : 100 QUAN :
R. R. T: 0. 87 MASS : 101. AMNT :
MASS INTENSITY MASS INTENSITY MASS INTENSITY
47 111 101 1000 103 573
LIBRARY EDITOR DATA: LIBRARYU1:
C3. H4. 0 M. WT.: 56
NAME: ACROLEIN
F. T. : . 925 SWW : 120 QUAN :
R. R. T: 0. 88 MASS : 56. AMNT :
MASS INTENSITY MASS INTENSITY MASS INTENSITY
52 21 54 31 56 1000
53 115 55 807
LIBRARY EDITOR DATA: LIBRARYU1:
C2. H2, CL2 M. WT.: 96
NAME: 1,1-DICHLOROETHENE
F. T. : 800 SWW : 100 QUAN :
R. R. T: 0. 91 MASS : 61. AMNT :
MASS INTENSITY MASS INTENSITY MASS INTENSITY
59 25 62 70 96 655
60 188 63 319 98 411
61 1000 70 11 100 50
LIBRARY EDITOR DATA: LIBRARYU1:
C3. H3. N M. WT.: 53
NAME: ACRYLONITRILE
F. T. : 950 SWW : 120 QUAN :
R. R. T: 0. 92 MASS : 53. AMNT :
A
100.00 • NG
MASS INTENSITY
105 66
11 BASE M/E:
RET. T:
A
500.00 NG
MASS INTENSITY
12 BASE M/E:
RET. T:
A
100.00 NG
MASS INTENSITY
13 BASE M/E:
RET. T:
A
500.00 NG

56
9:23

61
9:35

53
9:44
MASS INTENSITY
  52   684
MASS INTENSITY
  53  1000
MASS INTENSITY   MASS INTENSITY

-------
                                                              Section No.  XI
                                                              Revision No.  0
                                                              Date:  08/19/SI"
                                                              Page   6 of  11"
                  Tables of Libraries XI-1.  (Continued).
LIBRARY EDITOR
C  H2  CL2
NAME:      METHYLENE CHLORIDE
F. T.  :        950  SWW  :
R. R. T:      0. 92  MASS :
                               DATA:  LIBRARYU1:
                               M. WT.:   84
                                 200
                                  86.
          QUAN
          AMNT
MASS INTENSITY
  82    11
  84  1000
                 MASS INTENSITY
                   85    32
                   86   638
   MASS INTENSITY
    87     10
    88     99
LIBRARY EDITOR                 DATA:  LIBRARYU1:
C2. H2. CL2                    M. WT.:   96
NAME:      TRANS-1.2-DICHLOROETHENE
F. T.  :        850  SWW  :       80     QUAN   :
R. R. T:      0. 95  MASS :       96.    AMNT   :
                 MASS INTENSITY
                   62    80
                   63   349
                   96  1000
MASS INTENSITY
  59    53
  60   282
  61   930
      >
LIBRARY EDITOR
C2. H4. CL2
NAME:      1.1-DICHLOROETHANE
F. T.  :        950  SWW   :
R. R. T:      0. 96  MASS  :
   MASS INTENSITY
    97     57
    98    579
    99     17

DATA:  LIBRARYU1:
M. WT.:   98
                                  60
                                  63.
          QUAN
          AMNT
MASS INTENSITY
  61    55
  63  1000
                 MASS INTENSITY
                   65   284
                   83   134
LIBRARY EDITOR
C. H. CL3
NAME:      CHLOROFORM
F. T.  :        900  SWW   :
R. R. T:      1. 00  MASS  :
   MASS INTENSITY
    85     80
    98     70

DATA:  LIBRARYU1:
M. WT.:  118
                                  35
                                  85.
          QUAN
          AMNT
MASS INTENSITY
  47   240
  48   146
  51   176
                 MASS INTENSITY
                   83  1000
                   84    55
                   85   631
   MASS INTENSITY
    87     96
   117     14
   118     23
  14


       A
  100.00   NG

MASS INTENSITY



  15
                                   BASE M/E:
                                   RET. T:
                          84
                        9:45
                                                                  BASE M/E:    96
                                                                  RET. T:   10:00
                                                          A
                                                     100.00
                               NG
MASS INTENSITY
100     70
                                                     16
       A
  100.00
               BASE M/E:    63
               RET. T:   10:08
         NG
MASS INTENSITY
100     40
  17
     A
100.00
                                                                  BASE M/E:    83
                                                                  RET. T:   10:27
           NG
MASS INTENSITY
120     22

-------
                                                              Section No.  XI
                                                              Revision No.  0
                                                              Date:  08/19/83
                                                              Page   7 of  11
                  Tables of Libraries XI-.1.  (Continued).
LIBRARY EDITOR                 DATA:  LIBRARYU1:
C2. H3. CL3                 '  M. WT.:  132
NAME:      1,1,1-TRICHLOROETHANE
F. T.  :        950  SWW  :  .'•     60     QUAN   :
R. R. T:      1. 03  MASS :  i    101;    AMNT   :
MASS INTENSITY
MASS INTENSITY
  97  1000
LIBRARY EDITOR
C2. H4. CL2
NAME:      1,2-DICHLOROETHANE
F. T.  :        875  SWW  :
R. R. T:      1. 03  MASS :
                                  MASS INTENSITY
                                  101     83
                                  117    146

                               DATA:  LIBRARYU1:
                               M. WT.:   98
                 50
                 98.
                                         QUAN
                                         AMNT
MASS INTENSITY
  49   210
  51    70
  61  ' 276

LIBRARY EDITOR
C. CL4
MASS INTENSITY
  62  1000
  63   268
  64   313
                                  MASS INTENSITY
                                   65     46
                                   98    185
                                  100     95

                               DATA:  LIBRARYU1:
                               M. WT.:  152
NAME:      CARBON TETRACHLORIDE
F. T.   :        850  SWW   :
R. R. T:      1. 00  MASS  :
                 60
                117.
                                         QUAN
                                         AMNT
MASS INTENSITY
  47   120
  49   132
MASS INTENSITY
 117  1000
 119   952
                                  MASS INTENSITY
                                  121    296
LIBRARY EDITOK
C6. H6
NAME:      BENZENE
F. T.  :        850  SWW   :
R. R. T:      1. 00  MASS  :
              DATA:  LIBRARYU1:
              M. WT.:   78
                 60
                 78.
                                         QUAN
                                         AMNT
MASS INTENSITY
  49    37
  50   150
  51   166
MASS INTENSITY
  52   213
  74    49
  76    65
                                  MASS  INTENSITY
                                    77     268
                                    78    1000
                                    79      72
                                                     18
                                         A
                                    100.40
                                                 BASE M/E:    97
                                                 RET. T:   10:44
                                                              NG
MASS INTENSITY
119    130
121     36

  19
                                                 BASE M/E:    62
                                                 RET. T:   10:47
     A
100.00
           NG
MASS INTENSITY
102     16
  20


       A
  100.00   NG

MASS INTENSITY



  21


       A
  100.00   NG

MASS INTENSITY
207      0
219      0
264      0
                                                 BASE M/E:   117
                                                 RET. T:   10:57
                                                                  BASE M/E:    78
                                                                  RET. T:   10:57

-------
                                                              Section No.   XI
                                                              Revision No/HF
                                                              Date:   08/19/85"
                                                              Page   8 of   11'
                  Tables of Libraries XI-1.  (Continued).
LIBRARY EDITOR
C3. H6. CL2
                               DATA:  LIBRARYU1:
                               M. WT.:   112
NAME:      1,2-DICHLOROPROPANE
F. T.  :        875  SWW  :       69     QUAN
R. R. T:      0. 39  MASS :       63.    AMNT
MASS INTENSITY   MASS INTENSITY
  61   276         63   969
  62  1000         64   252

LIBRARY EDITOR
C2* H. CL3
NAME:*     TRICHLOROEHENE
F. T.  :        850  SWW  :
R. R. T:      1. 00  MASS :
                                  MASS INTENSITY
                                   65    366
                                   76    548

                               DATA:  LIBRARYU1:
                               M. WT.:  130
                                  60
                                 130.
                        QUAN
                        AMNT
MASS INTENSITY
  60   456
  62   775

LIBRARY EDITOR
C. H. CL2. BR
                 MASS INTENSITY   MASS INTENSITY
                   95   982       130   1000
                   97   676       132    977
NAME:      BROMODICHLOROMETHANE
F. T.  :        875  SWW  :
R. R. T:      1. 05  MASS :
DATA:  LIBRARYU1:
M. WT.:  162
                                  70
                                  83.
                                         QUAN
                                         AMNT
MASS INTENSITY
  47   145
  48   102
  79    81
MASS INTENSITY
  81    39
  83  1000
  85   577
   MASS INTENSITY
    87    127
    91     32
    93     33
LIBRARY EDITOR                 DATA:  LIBRARYU1:
C4. H7. 0. CL                  M. WT.:  106
NAME:      2-CHLOROETHYLVINYL ETHER
F. T.   :        850  SWW   :       70     QUAN  :
R. R. T:      1. 00  MASS  :       63.    AMNT  :
MASS INTENSITY
  57   173
  62   214
  63  1000
                 MASS INTENSITY
                   64    94
                   65   305
                   71   108
                 MASS INTENSITY
                 105     13
                 106    344
                 108    105
                                    22
                                   BASE M/E:    62
                                   RET. T:    11:26
                                    100.00   NG

                                  MASS INTENSITY
                                  112     49
                                    23
                           A
                      100.00
                                   BASE M/E:   130
                                   RET. T:   11:26
                                                              NG
                                  MASS INTENSITY
                                  134    305
                                                     24
                                         A
                                    100.00
                                                 BASE M/E:     83
                                                 RET. T:    11:34
                               NG
                                                   MASS INTENSITY
                                                   127     81
                                                   129    113
                                                   131

                                                     25
                                                           23
                                                                  BASE M/E:    63
                                                                  RET. T:   11:50
                           A
                      200.00   NG

                    MASS INTENSITY

-------
                                                              Section No.  XI
                                                              Revision No.~TT
                                                              Date:  08/19/55"
                                                              Page   9 of  11
                  Tables of Libraries XI-1.  (Continued).
LIBRARY EDITOR                 DATA:  LIBRARYU1:
C3. H4. CL2                    M. VT.:  110
NAME:      CIS-1.3-DICHLOROPROPENE
F. T.  :        900  SWW  :       70     QUAN  :
R. R. T:      0. 98  MASS :      110.    AMNT  :
MASS INTENSITY
  49/  170
  51    71
MASS INTENSITY
  75  1000
  77   293
                                  MASS INTENSITY
                                  110    202
                                  112    127
LIBRARY EDITOR                 DATA:  LIBRARYU1:
C3. H4. CL2                    M. WT.:  110
NAME:      TRANS-1.3-DICHLOROPROPENE
F. T.  :        950  SWW  :       55     QUAN  :
R. R. T:      1. 00  MASS :      110.    AMNT  :
MASS INTENSITY
  49   182
  75  1000
MASS INTENSITY
  77   289
 110   226
                                  MASS INTENSITY
                                  112    151
LIBRARY EDITOR
C7. H8
NAME:      TOLUENE
F. T.  :        950  SWW  :
R. R. T:      1. 00  MASS :
                               DATA:  LIBRARYU1:
                               M. WT.:   92
                                  40
                                  65.
                        QUAN
                        AMNT
MASS INTENSITY
  50    34
  51    48
  62    28
  63    58
                 MASS INTENSITY
                   64    15
                   65   100
                   73     2
                   74     8  .
MASS INTENSITY
 77     10
 89     41
 91   1000
 92    602
LIBRARY EDITOR                 DATA:  LIBRARYU1:
C2. H3. CL3                    M. WT.:  132
NAME:      1,1.2-TRICHLOROETHANE
F. T.   :        900  SWW   :       40     QUAN   :
R. R. T:      1. 00  MASS  :       97.    AMNT   :
MASS INTENSITY
  61   579
  83   901
MASS INTENSITY
  96   345
  97  1000
                                  MASS INTENSITY
                                  101     45
                                  132    115
                                  134    102
                   26


                        A
                   100.00.   NG

                 MASS INTENSITY



                   27


                     *   A
                   100.00   NG

                 MASS INTENSITY



                   28
                                                 BASE M/E:    75
                                                 RET. T:   11:57
                                                 BASE M/E:    75
                                                 RET. T:   12:18
                                                 BASE M/E:    91
                                                 RET. T:   12:21
                                                          A
                                                     100.00
                            NG
                                  MASS INTENSITY
                                   93     44
                                    29
                                         A
                                    100.00
                                                                  BASE M/E:    97
                                                                  RET. T:   12:27
                                                              NG
                 MASS INTENSITY
                 136     21

-------
                                                              Section No.   XI
                                                              Revision No.~HCT
                                                              Date:  08/19/85"
                                                              Page  ID of   ll~
                  Tables of Libraries XI-1.  (Continued).
LIBRARY EDITOR
C. H. CL. BR2
              DATA:  LIBRARYU1:
              N. WT.:  296
NAME: DIBROMOCHLOROMETHANE
F. T. : 850 SWW : ,
R. R. T: 1. 03 MASS :
MASS INTENSITY
79 305
91 808
93 170
MASS INTENSITY
94 102
127 1000 '
129 901
LIBRARY EDITOR
C2. CL4
NAME: TETRACHLOROETHENE -
F. T. : 900 SWW :
R. R. T: 1. 04 MASS :
MASS INTENSITY
82 108
94 579
96 383
MASS INTENSITY
129 761
131 725
133 228
LIBRARY EDITOR
C6. H5. CL
NAME: CHLOROBENZENE
F. T. : 900 SWW :
R. R. T: 1. 08 MASS :
70
127.
MASS
131
208
210
DATA:
M. WT.:
70
166.
MASS
164
166
168
DATA:
M. WT.:
60
114.
QUAN :
AMNT :
INTENSITY
171
198
64
LIBRARYU1:
164
QUAN :
AMNT :
INTENSITY
767
1000
465
LIBRARYU1:
112
QUAN :
AMNT :
MASS INTENSITY
  50   488
  51   527
MASS INTENSITY
  56    68
  76    93
LIBRARY EDITOR
C8. H10
NAME:      ETHYLBENZENE
F. T.  :        900  SWW   :
R. R. T:      1. 09  MASS  :
   MASS INTENSITY
    77    986
    78     67

DATA:  UBRARYU1:
M. WT.:  106
                160
                 79.
          QUAN
          AMNT
MASS INTENSITY
  50    73
  51   171
  63    65
  65   141
MASS INTENSITY
  74    19
  77    97
  78    98
  79    48
   MASS INTENSITY
    89     28
    91   1000
    92     77
   105     39
  30


       A
  100.00   NG

MASS INTENSITY
                                                     31
                                   BASE M/E:    127
                                   RET. T:    12:57
                                                 BASE M/E:   166
                                                 RET. T:   13:05
                                                          A
                                                     100.00
                                             NG
                                                   MASS INTENSITY
                                                   170     91
                                                     32
                                                 BASE M/E:   112
                                                 RET. T:   13:49
  100.00   NG

MASS INTENSITY
112   1000
114    322

  33
                                                 BASE M/E:    91
                                                 RET. T:   14:03
  100.00   NG

MASS INTENSITY
106    251

-------
                                                              Section No.  XI
                                                              Revision No.  D"~
                                                              Date:  08/19/55"
                                                              Page  11 of  11
                  Tables of Libraries XI-1.  (Continued).
LIBRARY EDITOR
C. H. BR3
NAME:      BROMOFORM
F. T.  :        850  SWW   :
R. R. T:      1. 12  MASS  :
              DATA:  LIBRARYU1:
              M. HT.:  250
                100
                173.
       QUAN
       AMNT
MASS INTENSITY
  79    98
  81   103
  91   105
MASS INTENSITY
  93   140
  94    80
 171   510
MASS INTENSITY
173   1000
175    493
250     37
LIBRARY EDITOR                 DATA:  LIBRARYU1:
C2. H2. CL4                    M. WT.:  166
NAME:      1,1,2,2-TETRACHLOROETHANE
F. T.  :        925  SWW   :      120     QUAN   :
R. R. T:      1. 16  MASS  :       96.    AMNT   :
MASS INTENSITY
  60   157
  61   ,170
  83  1000
  85   637
MASS INTENSITY
  87    99
  95   151
  96   108
  97   100
MASS INTENSITY
 98     64
131     70
132     25
133     66
                   34
     A
100.00
               BASE M/E:   173
               RET. T:   14:40
           NG
MASS INTENSITY
252    126
254    115
256     31

  35
                                                 BASE M/E:    83
                                                 RET. T:   15:08
  100.00   NG

MASS INTENSITY
166     43.
168     55
170     25

-------
                                                             Section No.   XII
                                                             Revision No.5~"
                                                             Date:  08/19/55"
                                                             Page  1  of   2
                            IX.  Quality Assurance
A.   Spikes - A minimum of 1/10 of all samples prepared (1/10 which 1/10 fish,
     1/10 waters) will be prepared and stored (see Section V).  Sample spikes
     will be obtained from the OWA operator and injected onto the solid sample
     as in Section _ .  Water samples will be spiked by the operator with the
     addition of internal standards.

B.   Duplicate Spikes - A minimum of 1/10 of all samples prepared (1/10 solids,
     1/10 fish, 1/10 waters) will be replicates and will be prepared as in the
     spike preparation.
       *
C.   Travel blanks (water) will be supplied when practical from random sampling
     sites and continue to be recorded as a function of sample site, transpor-
     tation source, and travel time.

D.   Spiked travel blanks (water) will be sent to sample sources to be returned
     with the collected samples.

     Note:  Trave1 blanks should be sent pretuned by different means to identify
     best approach and the travel blanks, spikes sent to collection source
     accordingly,                                                          :

E.   Cross :noc .. ,,'ill be supplied and completed.

-------
                                                             Section  No.   XII
                                                             Revision No.  0
                                                             Date:  08/19/55"
                                                             Page  2   of   2"
F.   Limits of detection will be confirmed periodically by quantitation of
     detection limits by decreasing concentration of spikes until  compound is
     not found or cannot be confirmed by forward library searches.   The limits
     of detection should be calculated for each media type.

     All detection limits found will be recorded in a QA/QC log as well as
     blanks, travel blanks, spike recoveries, and travel blank recoveries.
     These values will be addressed when reporting data are media  types.

-------
                                                          \
                                                             Section No. XIII
                                                             Revision  No.*~TT
                                                             Date:  08/19/51"
                                                             Page  I  of   37 "
                               XIII.   MAINTENANCE


A.   To assure continuous proper functioning of the Finnlgan OWA, maintenance

     of the system should be performed as per Table XIII-1.  (from Finnlgan  OWA

     1000 Series GC/MS Service Manual  pp. 2.4-2.6).



B.   The following procedure should be utilized to clean the source and  rods

     when so indicated in Table XIII-1.



     The following procedure covers cleaning the ion source  and changing the

     electron multiplier.  Routine replacement of the filament assembly  1s

     recommended with each cleaning of the ion source, as it is not practical

     to clean the filament assembly.   You will need the following tools  and

     equipment.
          1/2-in. (13 mm) end wrench
          100-mm section of 5-in. (130 mm) 00 plastic pipe
          disposable nylon gloves
          lint-free industrial tissues
          long-nose pliers
          small screwdriver (l/8-in./3-mm blade)
          number 600 aluminum oxide abrasive
          cotton swabs
          ultrasonic cleaner
          methanol
          aqua regia
          stainless steel wire
          Bunsen burner or propane torch
          alignment fixture for ionizer assembly (base P/N 01504-20170;  pin
           P/N 01504-20180)                                               •
          replacement filament assembly (P/N 30004-60020)
          plastic bag
          volt-ohm-mi11iammeter
          dental inspection mirror

-------
                                                             Section  No. XIII
                                                             Revision No.  0
                                                             Date:  08/19/18"
                                                             Page   2 of  37 "
     You must have a clean and clutter-free place to work.
                                                                        f
     Removing the analyzer.  You will  need a l/2-1n. (13-mm)  end  wrench  or  an
     adjustable wrench and something to set the analyzer on when  it  1s removed.
     A short (100-mm) length of 5-1n.  (130-mm)  OD plastic "pipe works well.

     1.   Remove the cover from the mass spectrometer and set It  aside.

          Note:  Do not touch the inner surfaces of the vacuum manifold  and
          magnet well, flor the analyzer, with your bare hands.  Natural  body
          oils will contaminate and necessitate cleaning of any part that has
          been touched.  These oils will cause  long-term background  problems
          in the spectra if present.  Wear clean disposable vinyl gloves1
          when handling such parts.

     2.   Remove the magnet from the magnet well and set it aside, then  remove
          the magnet well.  See Figure XIII-1.

     3.   Make sure that the cables connected to the analyzer are properly
          identifed (n.b. RF1 and RF2), the disconnect the cables.
                           )
          Caution:  The transfer line  from the  glass jet separator extends  into
          the ion source (see Figure XIII-2).  It is necessary to swing  the far
1P/N 00301-03600 small; 00301-09700 medium.

-------
                                                       Section No. XIII
                                                       Revision No.~lT~
                                                       Date:  08/19/|f"
                                                       Page   3 of  37~
        Figure XIII-1.   Mass spectrometer with cover removed.




     end of the analyzer to the left (away  from  the GC) to clear the ion
  »
     source before withdrawing the analyzer from the vacuum manifold.  It

     helps to have somebody watching from the front.



4.   Remove the right,  left, and top capscrews that secure the analyzer

     the manifold, and  loosen the bottom one a little.  Hold the analyzer

     in place as you remove the fourth  screw to  prevent the analyzer from

     slipping and being damaged.



5.   Swing the analyzer to the left to  clear the ion source, and carefully

     withdraw it from the vacuum manifold.
6.   Place the analyzer on a suitable support  (Figure  XIII-3).  Place wads

-------
                                                           Section No.  XIII
                                                           Revision No.   0
                                                           Date:   08/19/83
                                                           Page    4 of   37 "
                        TRANSFER
                           LINE
Figure XIII-2.  Removing the analyzer from the vacuum manifold (plain view).
 Analyzer must be swung to the right to clear the interface transfer line.
        of clean, lint-free tissues in the ends of the vacuum manifold to
        keep dust out.

        Note:  Sterile technique must be used in the following steps.   Do not
        touch with bare hands any part that you do not intend to clean.  Wear
        disposable nylon gloves when handling clean parts.  Tools must be
        clean and free from oils.
        General considerations.  It is necessary to hold binding posts with
        pliers to keep them from rotating when you are loosening or tightening
        screws.  This is especially important with the connections to the
        quadrupole rods, where rotation of the binding posts could throw;the
        rods out of alignment; and also with the filament connections, where
        rotation of the posts could break the filament.

-------
                                                  Section No. XIII
                                                  Revision No.~^T~
                                                  Date:  08/19/55"
                                                  Page   5 of  37~
               Figure XII1-3.   Mass  analyzer.
Ion source:  Note the orientation of the ion  source assembly with
respect to the quadrupole housing before you  remove it.  The odds are
against you if you don't.  Disconnect the  leads  from the ion source,
then remove the three screws  that secure the  ion source assembly to
the quadruple housing and lift the ion source assembly from the
housing.  See Figure XIII-4 for identification of parts.

-------
                                                    Section  No.  XIII
                                                    Revision No.  0
                                                    Date:  08/19/83"
                                                    Page   6 of   3f"
                    ..  •  t ,  J /— ION VOLUME
                                       COLLECTOR
                                     LENS
                                     APERTURE
                           FILAMENT
                           ASSEMBLY
        Figure XIII-4.  Exploded view of ion source.
1.   Remove and discard the filament assembly.
2.   Remove the collector.
3.   Remove the ion volume, extractor, lens, and aperture.

-------
                                                   Section No. XIII
                                                   Revision No.~lT
                                                   Date:  08/l°/83
                                                   Page   1 of  37"
4.   Using a slurry of number 600 aluminum oxide in water and  a
     cotton swab, remove all  contamination from the ion  volume.
     (Contamination appears as dark  or discolored areas.)  Use a
     wooden stick cut at an angle to clean the Inside corners. Try
     not to let the slurry dry on the metal,  as dried aluminum oxide
     is difficult to remove.   Rinse  off the aluminum oxide with
     running water and place the parts in a small container of
     detergent and water.  Clean the extractor, lens, and aperture
     in the same way.

5.   Place the container of parts in an ultrasonic cleaner for 60
     seconds, then rinse with water.  From this point on, parts must
     be handled with clean tools to  prevent recontami nation.

6.   Using clean tools, place the parts in a container of water;  and
     place the container in the ultrasonic cleaner for 60 seconds.

     Caution:  Ultrasound may cause  the small screws in  the binding
     posts to vibrate loose.  Take care that none are lost in  rinsing.

7.   Rinse the parts with methanol.

8.   Place the parts in a clean beaker of methanol, and  place  the
     beaker in the ultrasonic cleaner for 60 seconds.

-------
                                                   Section No. XIII
                                                   Revision No,  0
                                                   Date:  08/19/ST
                                                   Page   8 of  37"
9.   Inspect the parts,  and ensure  that  all  spots  and particles  have
     been removed.

10.  Immerse the sapphire spacers 1n aqua  regia  and  rinse  them with
     distilled water.                           -

11.  If the ceramic parts have metal deposits  (dark  streaks)  on  them,
     immerse them in aqua regia and rinse  them with  distilled water.
     Remove organic contaminants by stringing  the  parts  on a  stainless
     steel wire and heating them in a Bunsen or  propane  flame until
     they are red hot.

12.  Check to ensure that each of the threaded rods  extends 13 mm
     (1/2 in.) above the ionizer adapter,  then set the adapter on the
     alignment fixture base (Figure XIII-5).

13.  Stack the aperture, lens, extractor,  and  ion  volume assembly on
     the studs, using the appropriate bushings and spacers (see
     Figure XIII-4).  Install the four nuts, but do  not  tighten  them.

14.  Insert the three screws into the mounting holes for the filament
     assembly and collector.  Leave the screws loose for later removal

15.  Insert the alignment-fixture pin through  the  ion volume, lens,

-------
                                     PIN
                                     01504-20180
                                      ION
                                      VOLUME


                                      EXTRACTOR

                                      LENS


                                      APERTURE

                                      ADAPTER
                                      BASE
                                      01504-20170
                               STORE PIN HERE
                               WHEN NOT IN USE
                                                    Section No. XIII
                                                    Revision No.  iQT
                                                    Date:  08/19/8T"
                                                    Page   9 of  37 "
         Figure XIII-5.  Alignment  fixture base and
              pin, with  assembled  ion source.
     extractor, and aperture and  into the alignment-fixture base.



16.  Tighten the four  nuts.   (DO  NOT OVERTIGHTEN; finger-tight is

     enough) and remove the alignment-fixture pin.
17.  Remove the three  loose  screws  and install  the new filament :

     assembly.  See Figure XIII-6.   Position tf? f r-..nent assembly

-------
                                                   Section  No.  XIII
                                                   Revision No.  0
                                                   Date:  08/19/83
                                                   Page   10 of   37"
     parallel and as close as possible to the Ion volume without

     touching.
       Figure XI11-6.  Ion source with filament assembly
          installed.  Note arrangement of spacers.
18.  Look through the ion volume from the collector side and be sure

     that the filament is visible in the center of the slit (Figure

     XI11-7).  Tighten the screws.
        Figure XIII-7.  Ion source assemb.y pr,-.
          installation o.  collector.  Loc  thr-'
           the ion volume from the col lee or $v
             and make S1."-? that the filar  '
             visible in the center of the slit.

-------
                                                   Section No. XIII
                                                   Revision No.  0
                                                   Date:  08/19/83T
                                                   Page   11 of  37'
19.  Install the collector,  making sure that  the bushing  and  spacers
     are 1n the proper sequence (Figure XI11-8).
            Figure XIII-8.  Ion source assembly
              with collector installed.  Note
                  arrangement of spacers.
20.  Position the collector so that it is parallel to and 0.5 to 1.0
     mm from the ion volume, then tighten the screw.
     Quadrupole:  Normally, contamination occurs only at the ends of
     the rods nearest the ion source.  If the rods are dirty, discon-
     nect the rf leads and unscrew the four binding posts.  Note:  if
     the ends of the rods are marked with a punch mark (characters or
     center punch).  It will be necessary to reinstall the rods with
     the punch-marked ends facing in the same direction after they
     are cleaned.  Remove the single screw holding the quadrupole

-------
                                              Section No.  XIII
                                              Revision No.  0
                                              Date:   08/19/85"
                                              Page  12 of   37
assembly to the housing and, grasping one rod, draw the assembly
from the housing.  Handle the quadrupole assembly by the ceramic
Insulators to avoid misalignment.  Place a clean plastic bag
over the remainder of the analyzer to keep dust out of the
multiplier.  Using number 600 aluminum oxide/water slurry as
before, clean the quadrupole rods.  Do not disassemble the rods
from the ceramic spacers.

Caution.  To avoid stressing the assembly and possible misalign-
ment, rest each rod against a stable support while cleaning.

Rinse the ends of the rods with methanol.  Do not allow methano'l
to come into contact with the ceramic insulators.  The ceramic
will absorb methanol and generate a strong methanol background
when the filter is put back into service.  Slide the filter
assembly into the housing, making sure that the marked ends of
the rods are facing the same wasy as before.  Replace and tighten
the screw that secures the filter assembly to the housing.
Replace the four binding posts.

Caution.  Tighten the binding post finger tight only.  Over-
tightening may turn the rod-retaining screws, necessitating.
realignment of the rod assembly.

-------
                                                   Section No. XIII
                                                   Revision No.  0
                                                   Date:  08/19/53"
                                                   Page  13 of  37
     If the electron multiplier Is  not to be changed, proceed to
     Reassembly.

     Electron multiplier.   The normal  life  of  an  electron multiplier
     ————                ^
     is 6 to 18 months.  We recommend that  you budget for a new
     multiplier once per year.  To  change the  electron multiplier,
     proceed as follows.

1.   Disconnect the leads  from the  ion source  and quadrupole filter.

2.   Remove the three screws at the base of the multiplier housing
     and lift off the entire assembly, exposing the multiplier.

3.   Loosen the upper screws (Figure XIII-9) in the high-voltage and
     anode couplings and remove the two  screws that secure the
     multiplier to the flange plate.  Remove the  multiplier.

4.   Install the  new multiplier on  the flange, making sure that the
     leads are connected to the proper terminals  (bottom) lead to
     ANODE, upper lead to HV).

5.   Replace the  housing,  aligning  the punch mark on the housing.
     with the punch mark on the flange (Figure XIII-10).

-------
                                             Section No. XIII
                                             Revision No.  0
                                             Date:  08/19/5T"
                                             Page  14 of  37
      Figure XIII-9.  Electron multiplier.
      Loosen screws a and remove screws b
      to replace the electron multiplier.
               TAPPED HOLE
                         PUNCH MARK
                    PUNCH MARK
Figure XIII-10.  Closeup of Multiplier housing,
              showing punch marks.

-------
                                                   Section No. XIII
                                                   Revision No.  0
                                                   Date:  08/19/33"
                                                   Page  15 of  37 '
     Reassembly:



1.   Reconnect the rf leads to the quadrupole mass  filter.



2.   Reinstall the ion source and connect the leads.



3.   Using an ohmmeter, check that no element is  shorted to any other

     or to ground.
     RF1 to RF2
     Ground to RF1
     Ground to RF2
     Ground to lens
     Ground to extractor
     Ground to ion volume
     Ground to collector
     Ground to filament 1
     Ground to filament 2
     Lens to extractor
     Lens to ion volume
     Lens to collector
     Lens to filament 2
     Lens to filament 2
     Extractor to ion volume
     Extractor to collector
     Extractor to filament 1
     Extractor to filament 2
     Ion volume to collector
     Ion volume to filament 1
     Ion volume to filament 2
     Collector to filament 1
     Collector to filament 2
     Filament 1 to filament 2
\-~/
(-0)
(-0)
(-0)

131
open circuit (»n)
open circuit («n)
open circuit (-0)
open circuit
open circuit
open circuit
open circuit
open circuit , „,
open circuit («8)
open circuit («8)
open circuit («n)
open circuit («n)
open circuit (-Q)
open circuit («n)
open circuit (•£})
open circuit (•«)
open circuit
open circuit
open circuit
open circuit
open circuit  *fi
open circuit (•«)
open circuit («8)
         0.2 ohms
(»n)
(•«)
(»n)
     Installation:

-------
                                                  Section No.  XIII
                                                  Revision No.  0
                                                  Date:  08/19/55"
                                                  Page  16 of  37
     Caution.   The transfer line from the glass jet separator projects
     Into the  1on source  when  the analyzer 1s Installed.  Angle the
     analyzer  well to  the left (as viewed from the rear) and maintain
     this attitude while  Inserting It into the vacuum manifold, until
     the flanges meet.  Have someone watch from the front to make
     sure that the transfer  line enters  the  ion source without damage.

1.   Install the analyzer, securing  1t with  the four screws.  Tighten
     the screws.  Tighten the  screws uniformly.

     Note:  It is good practice to tighten the  screws a  little at  a
     time in diametrically opposite  pairs.   See Figure XIII-11.
                            0
                            $
                              *
                              *
                            >  •
        Figure  XIII-11.  Rear view of vacuum manifold.
                Tighten  screws  in this order:
                1,  3,  2, 4,  1,  3, 2. 4, etc.

-------
                                                   Section No. XIII
                                                   Revision No.  0
                                                   Date:  08/19/31"
                                                   Page   17 of  37"
2.   Using a dental Inspection mirror, check that the transfer line
     Is 1.5 to 3 mm from the 1on volume.  Reposition  the  source,  if
     necessary, then check-that the transfer line 1s  still  properly
     aligned with the Ion-source opening.

     Note:  If the Instrument Is equipped with a solid-probe  Inlet,
     the opening In the Ion volume must be accurately centered with
     respect to the Inlet barrel.  If 1t Is not, loosen the three
     screws that secure the Inlet assembly to the vacuum  manifold and
     shift the Inlet assembly.

3.   Use the VOM to check that none of the ion-source components  is
     shorted to any other, then attach the cables to  the  rear
     connectors.

4.   Install the magnet well and tighten the.four capscrews.
                                                             «

5.   Move the PUMP switch on the power controller to  PUMP and the
     ANALYZER VOLTAGES switch to ENABLED.  Wait 10 minutes.  Press
     the yellow RESET button.  The RESET light should go  out; if  not,
     check the SYSTEM STATUS lights.

60   vmall the magnet in the magnet well in a vertical  orientation
     aiM connect the VOM to test points 6 (+) and 7 (-} on the

-------
                                                       Section No. XIII
                                                       Revision No."~TT
                                                       Date:  08/19/1F
                                                       Page  18 of  37
          Ion-source board  (Figure XIII-12).   (Test points are numbered
          from top to bottom of the  board.)
                       /.     "
     7.    Start the data'system and  run the manual tune program (MT).
          Turn on the filament.  The VOM  should  indicate between 0 and 250
                        .t
          mV DC.  Rotate the magnet  back  and forth and move It in and out
          while watching the VOM, the object being to obtain maximal
          collector current.  (The measurement is across 100 ohms, so 100
          mV represents 1 mA of collector current.)  If the maximal
          reading is less than  60 mV, turn the magnet end for end and try
          again.  Low collector current may  indicate that the filament is
          not properly aligned.

          Note:  Collector current  should be 80% of the total emission
          current, which can be read at test points 6  (-) and 8 (+).
          Emission current"is factory set at 750 pA  (read 75 mV) and
          should require no attention.

     8.    Turn the ANALYZER VOLTAGES switch  to STANDBY  (or run the  system).

Vacuum System:

Purging the Alcatel pumps:   The Alcatel  vacuum pumps can be purged  without
interrupting the system operation.   Unscrew  the  knob shown in Figure XIH-13

-------
                                                      Section No.  XIII
                                                      Revision Itoo  0 ,
                                                      Date:   08/19/%3"~
                                                      Page _19 of   37
Figure Xiil=12.  Ion-source circuit boardo  Connect VOM to TP 4(+)
         and TP5 (-).  Test points are numbered from top.

-------
                                                        Section No.  XIII
                                                        Revision No.  0
                                                        Date:  08/19/55"
                                                        Page  20 of  37
                             GAS
                            BALLAST

                     SIGHT
                     GLASS
 GAS
BALLAST
 VALVE
                              DRAIN
                              PLUGS
              Figure XI11-13.  Alcatel pump, showing gas
                   ballast valve, drain plug, fill
                       plug, and sight glass.
a few turns to open the gas ballast valve.  Let the pump run with the gas

ballast valve open for about 5 minutes, then close the valve.



Changing the oil in the Alcatel pumps.  Oil changes should  be  performed

while the pump is warm.  Shut down and vent the system.  Disconnect the

pump hose and power cord.  Place the pump  on a bench.  Remove  the drain

plug Figure XIII-13) and allow the oil to  drain,  the  replace the plug and

fill the pump with fresh oil to the middle of the sight  glass. Alcatel

oil VP1 is recommended.  Start up the system.
Changing the oil  in  the  turbomolecular pump.   To change the oil  in the

turbomolecular pump,  stop the pumps and vent  the vacuum system as

described  under Mass  Spectrometer  maintenance, above.   Then follow tilt-

instructions in the  legend to Figure  XIII-14o

-------
                                                          Section No. XIII
                                                          Revision No.  0
                                                          Date:  08/19/5T"
                                                          Page  21 of  37
                                      OIL RESERVOIR
Figure XIII-14.  Turbomolecular pump.   Before removing the oil  reservoirs,
  vent the vacuum system to the atmosphere, then unscrew the reservoirs,
   taking care not to damage the 0-rings.  Wipe out the reservoirs with
   e clean rag.  Wipe the 0-ring and oil it lightly with T12 oil.  Fill
   the reservoirs with 10 ml of T12 oil and reattach them to the pump.
                            (Tighten lightly.)
  Regenerating the filter/drier:  The filter/drier (Figure XIII-1) is filled

  with indicating DRIERITE (anhydrous calcium sulfate).  Indicating DRIERITE

  is a distinct blue.  When exhausted, it turns to a rose red or pink.

  Regeneration will restore the blue color.  DRIERITE granules should be

  removed from the col urn and spread evenly, one granule deep, on a tray.

  Heat for 1 h at about 200C.  The dessicant should then be cooled in a

  tightly-closed container before you refill the column.  Felt filters

  should be dried at 100°C for about 30 minutes before reassembly.
  DISK DRIVE:

-------
                                                        Section No. XIII
                                                        Revision No.T"
                                                        Date:  08/19/51"
                                                        Page   22 of  37
Caution.  There is no warranty on the read/write heads nor the  disks;
preventive maintenance of these items is the customer's responsibility.
It is therefore extremely important that the preventive maintenance
procedures suggested in this manual be performed at suitable  intervals.
We suggest inspection of the air filter and the read/write heads  three
months after the installation of the system and at regular intervals
thereafter, based on the condition of these items at the first  inspection.

Read/Write Heads:        .

Because of the very small clearance between the head and disk surfaces,  it
is,imperative that the heads as well as the disks be kept clean.   When in
the flying attitude, the heads are separated from disk surface  by only a
few micrometers; and because of the high disk speed, even small particles
adhering to the heads may cause damage to the head or to the  surface  of
the disk.  To inspect the heads for accumulation of dust or oxide parti-
cles, remove the disk pack and illuminate the head area.  The upper heads
can be seen with the aid of a dental-type mirror.

Caution.  Do not touch the head surfaces or bring them into contact with
any object other than the cleaning applicator.

To clean the heads, use 91% isopropyl alcohol in solution with  9% distilled

-------
                                                            Section No. XIII
                                                            Revision No.  0
                                                            Date:  08/19/31"
                                                            Page  23 of  37
     water.   Do not use any other  kind of  liquid.  Apply with a lint-free

     cotton  swab1.



     Spindle Chuck and Cone:



     Clean the surfaces of the spindle chuck  and cone  (Figure XIII-15) using

     alcohol-moistened disposable  wipes  (Kimwipes).  Ferrous particles adhering

     to the  magnet ring can be removed with masking tape.
           Figure XIII-15.   Disk drive showing  spindle  chuck  and  cone.
     Air Filter:
     To inspect the air filter, pull  the disk  drive out  on  its  rails  and  remove
IPerkin-Elmer recommends Foam Floss-tipped swabs,  which  leave no  residue.

-------
                                                        Section No.  XIII
                                                        Revision No.   0
                                                        Date:   08/19/83
                                                        Page  24 of   37
the sheet metal cover under the left front corner (seven screws,  see
Figure XII1-16).  Loosen the two wedge-shaped clamps on the filter and
remove the filter element.   Peel  off the styrofoam prefliter and  Inspect
the filter.  A clean filter is pure white.  A dirty filter will be white
around the edges where the adhesive for the prefliter blocks the  flow of
air and gray or black 1n the center.  If the filter is dirty, 1t  must be
replaced.
                                            FILTER
                  Figure X-16.   Air filter removal.

-------
                                                             Section No. XIII
                                                             Revision  No.""TT~
                                                             Date:  08/19/55"
                                                             Page   25  Of   37"
     Parts Order Requests:



     To order parts, call the Finnlgan Parts Center (408)  732-0940.



C.   The following troubleshooting charts (Tables XIII-1 thru  XIII-4)  should  be

     utilized if difficulties are encountered in OWA operation.   These charts

     should be used before contacting Finnigan Service Representative.



     Note:  Maintenance and troubleshooting guide (Tables  XIII-1  thru  XIII-4)

     from Finnigan manuals 10000-90080 and 10000-90140.

-------
                                                                               Section Wo0  am
                                                                               Revision Mo0   0
                                                                               Oitet   08/19/33
                                                                                      26 of   37
                  TABLE XIII-lo   MAINTENANCE  CHECKLIST (PREVENTATIVE  MAINTENANCE):
   1 ©tesa jc/j
   2. ©laso jc3
   3. fctoss onctyxcr ^^
                 assy
    o aotcr Rgc; scnssr swi
              moso
  g»
  6. Atesd vcs^atm giS5^3 (§8
  y.
  Q.
           SJ/SJCTfl
           e3
    °cga saseo BJcRtofa assy
   D. Ai7 sate? as feciferai e3 eago
   afco  • •
   3. S^rcJ eafeto ea
'   fl. Yapo
          Topo
;   fl. PoKi
f&eocisago
      loohs.
         oco.
                            fcyi?y goes sftapo.
 ©3 as?
                                     & ea
                                                         CT
                                                     B5C3C1
                                                            o
                                                               &
                                                                        EOTSQVAL
                                           eioa«.
                                                                            oo
                                                                            ore &?
                                     li

-------
                                                                         Section NOc ___
                                                                         Revision Wo<>   0
                                                                         Date:   08/19/15"
                                                                                27 of   37
      TABLE-XIII-1.  MAINTENANCE CHECKLIST (PREVENTATIVE MAINTENANCE)  (Continued),
 fl. Para
S. Afijustasio Sg
•3. P>e£)aenc? garage
  ratio
S.
 0. Aff fifsc?
 ?. UTO fusoa to jfto ©g
        tef packed
IL
        soptura in 610 ©6
i. Conic? gas
            Wtc?
 7.
 0.
 0. Pceted coluffm (glass
?@. Fceticd eehjma ((nets!
m. OcHcae? pe^ i© Cg-ftfl
  . ©6 e@o3 esc/a fcra
         goa
                                to
tho

-------
                                                             Section No.  XIII
                                                             Revision No.  0
                                                             Date:   08/19/83
                                                             Page   28 of   37
Item
         TABLE XI11-2:   ELECTRONIC TROUBLESHOOTING CHART
     s=css=s==Bssssssssasssss===ssssss=sss=sass=ssss=as8ssa8ssssssc=BS
     Symptom                Probable Cause           Corrective Action
               NOTE:  Unless otherwise specified,  all  paragraph
                      and page references pertain  to the Sigma 3
                      Service Manual  from Perkin-Elmer (part no.
                      00920-05902, August 1978 issue).
  1.    GC will not enable.
  2.
Oven fan will not
rotate.
  3.


  4.
No display on front
panel.

Oven,  injector and
detector will not
heat.
                     Fl blown
                     F2 blown
                     SI (power switch) open
                             Replace Fl
                             Replace F2
                             Check SI with ohm
                             meter.
Lower fan tangled in cables  Reposition cables.
                             Oven lid micro switch
                             defective
                             Kl oven relay defective.
                             B2 fan motor defective.
F4 blown +5VDC
Fl in power supply blown
F2 in power supply blown
F3 in power supply blown
F4 in power supply blown

+15VDC power supply failure
-15VDC power supply failure
+5VDC power supply failure
+12VDC power supply failure
                             MPU PCB defective
Check for continuity
with ohm meter; if
required, replace the
faulty switch.

Check for continuity
with ohm meter; if
required, replace the
faulty relay.

Check for proper fan
rotation and check
that windings are not
open.  If defective,
replace faulty motor.

Replace F4.
Check with ohm meter
and replace defective
fuse.
                                                  Check power supply
                                                  with volt meter and
                                                  replace.  If defec-
                                                  tive, refer to section
                                                  3.4.3.

                                                  Check TP14 (DAC vol-
                                                  tage); if incorrect,
                                                  replace MPU PCB card.

-------
                                                           Section No. XIII
                                                           Revision No.  6
                                                           Date:  08/19/8?"
                                                           Page  29 of  37 '
Item
           Symptom
TABLE XIII-2.  (Continued).

        Probable Cause
Corrective Action
5.    Oven will not heat   Oven max temp set too low
      but Injector and
      detector will heat.
                           Oven heater open
                           Defective control  circuit
                           on temp control PC8 card
                           No set point from MPU
                           Oven Triac defective
                           Oven lid micro switch open
6.     Injector will not    Defective control circuit
       heat oven  and        on the temp PCB card
       detector will heat.
                                                          Check oven max  1s set
                                                          correctly, power-down
                                                          and  power-up.   Reset
                                                          oven temp.

                                                          Check heater with ohm
                                                          meter;  if open,
                                                          replace hester.

                                                          Check TP7 (oven Triac
                                                          pulse); 1f incorrect,
                                                          check TP3, if correct,
                                                          replace temp
                                                          corrective action PCB
                                                          card; 1f Incorrect,
                                                          check oven sensor with
                                                          ohm  meter.

                                                          Check TP8 (oven set
                                                          point); if incorrect,
                                                          replace MPU  PCB card.

                                                          Check oven Triac wave-
                                                          form with a  suitable
                                                          scope.

                                                          Check switch for con-
                                                          tinuity with ohm
                                                          meter;  if defective,
                                                          replace the  switch.

                                                          Check TP9 injector
                                                          Triac  pulses;  if
                                                          correct, check
                                                          injector Triac
                                                          waveform; if
                                                          incorrect, check TP5.
                                                          If TP5  is correct,
                                                          replace temp PCB card;
                                                          if incorrect, check
                                                          injeccor sensor with
                                                          ohm meter.

-------
                                                            Section No. XIII
                                                            Revision No.  0
                                                            Date:  08/19/1?"
                                                            Page  30 of  37~
Item
               TABLE XIII-2.  .(Continued).
: = === ===== === ==== = = = = ======= ==3=:== ==== ==== =
Symptom               Probable Cause
Corrective Action
  6.                          Injector  heater cartridge
                             open


                             Injector  Triac defective
  7.    Detector will not    Defective  control circuit
        heat oven and        on temp PCB card
        injector will heat.
        Injector and         No set point  from MPU  PCB
        detector will not    card
        heat.
        Unable to read any   GC/IO PCB card defective
        actual temps or set
        points.
                             MPU PCB card defective
                                             Check for continuity
                                             with ohm meter and
                                             replace, if defective.

                                             Check injector Triac
                                             waveform; 1f
                                             incorrect, replace
                                             Triac.

                                             Check TP8 detector
                                             Triac pulses; 1f
                                             correct, check
                                             detector Triac
                                             circuit; if incorrect,
                                             check TP12, and if
                                             TP12 is correct,
                                             replace temp PCB card.
                                             If TP12 is incorrect,
                                             check injector sensor
                                             with meter.

                                             Check TP9 zone set
                                             point; if incorrect,
                                             replace MPU PCB; if
                                             correct, check for
                                             loose or open cable
                                             from MPU to temp
                                             control.

                                             Connect test box in
                                             place of GC/IO; if GC
                                             works, replace GC/IO;
                                             if GC does not work,
                                             check MPU.

                                             Check MPU TP14 (DAC
                                             voltage); if
                                             incorrect, replace MPU
                                             PCB card; if correct,
                                             check cable from
                                             keyboard to MPU.

-------
                                                             Section No. XIII
                                                             Revision No.~"TT~
                                                             Date:  08/19/53"
                                                             Page   31 of   37~
Item
                    TABLE XIII-2.  (Continued).
== = ==== = = = ======== = = === = =====================:======:r=:5== ==c======== ==r:r
     Symptom                Probable Cause           Corrective Action
  9.
  10.
  11.
  12.
  13.
Actual temps are
Incorrect.
                     Trip, Ready, and Display
                     PCB card defective
Temp control PCB card needs
alignment

MPU needs alignment
READY indicator      Trip, Ready, and Display
light will not come  PCB card defective or
on.                  misaligned.
                             MPU PCB card defective
Oven will not pro-   MPU PCB card defective
gram temperature.

Oven door solenoid   +24VDC missing
will not operate.
                             MPU PCB card defective
If MPU and keyboard
are working correctly,
replace Trip, Ready,
and Display PCB card.

Align temp PCB card as
per paragraph 4.12.1

Align MPU PCB card as
per paragraph 4.12.3.

Check part number of
IC4, refer to para-
graph 3.7.2b; If in-
correct, replace dis-
play PCB card; if
correct, check PIN1 of
IC13.

Check PIA2 pin 4 for a
logic high value.
Refer to paragraph
3.7.2B.  If P1A2 is
incorrect, replace MPU
PCB card.

Replace MPU PCB card.
                             Check +24VDC, if
                             faulty, replace power
                             supply PCB card.

                             Check PIA2 Pin 6 (a
                             high signal is an open
                             vent command); refer
                             to paragraph 3.7.3; if
                             incorrect, replace MPU
                             PCB card.

-------
                                                            Section No. XIII
                                                            Revision No.  0
                                                            Date:  08/19/83
                                                            Page 32  of  37"
Item
                TABLE XIII-2.   (Continued).
===========================================
 Symptom               Probable Cause
                                                    Corrective Action
  13.
  14.
Sub-ambient assem-
bly will not work.
                     Trip, Ready, and Display
                     PCB card defective
                             Clogged  flow restrictor in
                             air line
                             Faulty mini  piston
                 Sub-ambient  switch  in  OFF
                 position
                 CO2 Tank empty  or not  dip
                 tube type

                 Trip, Ready, and Display
                 PCB card defective
                             +24VDC power supply is
                             faulty
                             Pressure too low (LN2
                             only)
Check collector of
Q12; refer paragraph
•3.7.3; if incorrect,
replace display PCB;
if correct, check
solenoid with ohm
meter.

If vent solenoid is
actuating, the flow
restrictor should
allow piston to open
fully in 45 t 15
seconds; if not,
replace restrictor.

Disconnect air line at
piston, and check that
piston moves freely;
if the piston does not
move freely,, then
replace it.

Position the sub-ambi-
ent switch on back of
GC to ON.
                                                  Check  collecor  of  Q12;
                                                  refer  to paragraph
                                                  3.7.3; if incorrect,
                                                  replace display PCB
                                                  card;  if correct,
                                                  check  S/A solenoid.

                                                  Check  24VDC  power
                                                  supply, if power
                                                  supply is faulty,
                                                  replace power supply
                                                  PCB card.

                                                  Check  that pressure is
                                                  20 to  30 Ibs. at tank.


-------
                                                             Section No. XIII
                                                             Revision No.  13"
                                                             Date:  08/19/55"
                                                             Page  33 of  37
Item
 TABLE XII1-3.  CHROMATOGRAPHY TROUBLESHOOTING CHART
S==S2=S=32S=S=SS=SSS=S===8SS8SS=S==SSS2S===SSSS8S3:SSSSSCSS8SSSSSS

Symptom                Probable Cause           Corrective Action
                         NOTE:   Unless otherwise specified,  all
                         paragraph and page references pertain to
                         the Sigma 3 Service Manual  from Perkin-Elmer
                         (part no. 00920-05902, August 1978  issue).
  1.    Non reproducible
        quantitation
                Sample evaporation

                Septum leaking

                Faulty syringe

                Detector or injector too
                cool
                             Insufficient head pressure
                             on regulator or controller

                             Ghost peaks
                             Leaks

                             Oven temperature unstable
                             Defective pressure regula-
                             tor or flow controller
Use fresh sample.

Change septum.

Use test syringe.

Check with temp meter,
if the reading is
lower than set
temperature, refer to
the electronics
troubleshooting guide.

Increase input
pressure.

Contaminated septurns,
injector and column;
clean or replace
column.  Refer to
paragraph 4.3.2B.

Section 4.4.

Check with temp meter
and if incorrect,
refer to the
electronic trouble-
shooting guide.

Bypass to test.
                                                                    (continued)

-------
                                                           Section No.  XIII
                                                           Revision No.   0
                                                           Date:   08/19/37"
                                                           Page  34 of   37"
Item
Symptom
TABLE XI I 1-3.  (Continued).
sssaassss=as=3Ssssassssssss
        Probable Cause
                                                           Corrective Action
2.    Peak tailing
3.    Low response
 Injection too large

 Dead space in column
 adapters

 Incorrect temp for sample

 Column condition

 Injector, oven detector
 temp out of calibration
                           Incorrect column for
                           application
                                             Reduce volume.

                                            'Change fittings.


                                             Review application.

                                             Use a test column.

                                             Use temp meter.  If
                                             defective, refer to
                                             paragraph 4.12.1.

                                             Review application.
 Leaks at column connections  Refer to Section 4.4.

 Septum leak                  Replace septum.

 Faulty syringe               Use test syringe.

 Contaminated Gases           Change cylinders.
                           Poor injection technique
                              Review injection
                              procedures.

-------
                                                             Section No.  XIII
                                                             Revision No.   0
                                                             Date:   08/19/8T
                                                             Page  35 of   37
            TABLE XII1-4.  TROUBLESHOOTING CHART FOR ANALYZER MODULE
a==ss==rss=========s================================================
Item         Symptom                Probable Cause              Remedy
  1.    Low sensitivity
Ion source contamination.

Magnet misalignment.

Contaminated lens.
                             Defective or misaligned
                             quadrupole.
                             Contaminated rod.

                             RF failure.
Refer to operating
manual cleaning
procedure.

NOTE:  With an emis-
sion current of 750
yA from the filament.
the collector current
will be at least 600
yA.  The filament
alignment and magnet
adjustment will
influence this value,
and if these are
optimal then collector
currents below 600 yA
indicate a contamin-
ated source; if the
collector current is
below 400 yA the
source is grossly con-
taminated.  Another
indication of source
contamination is the
need for higher than
normal extractor
voltage; greater than
5 volts.

Refer to the
Operator's Manual
quadrupole alignment
procedure.

NOTE:  Quadrupole
filter contamination
is usually diagnosed
as differential loss
of high mass sensitiv-
ity, poor resolution
and poor peak shapes.
Together these factors
                                                                    (continued)

-------
                                                             Section No. XIII
                                                             Revision No.  0
                                                             Date:  08/19/%T
                                                             Page  36 of  37"
s===================r

Item         Symptom
TABLE XIII-4.  (Continued).
      :sss2ssss==s==s=ss^=s:
         Probable Cause
   Corrective Action
  1.
  2.    Low output
        Calibration
        difficulty
        No  spectra
  Multiplier gain decreased
  due to multiplier failure
  or contamination.          v

  Malfunction in high voltage
  circuit.
  Poorly tuned source or air
  leak.
                             Contaminated quadrupole
                             filter.
  Filament failue.

  No filament voltage.
cause progressive
difficulty  in tuning
•and calibration.  The
rods are contaminated
and should  be cleaned.
Refer to the
Operator's  Manual
maintenance procedure.

Replace multiplier.
 Troubleshoot  High
 Voltage  circuit  (see
 Test  procedures  and
 adjustments in Section
 10  of this manual).

 Adjust MT and/or AT,
 check for air leak as
 described in  Section
 13  of this manual.

 Clean rods according
 to  operating  manual
 maintenance procedure.

 Replace  filament.

 Check for filament
 voltage  at the  fila-
 ment  connector.
 Troubleshoot  power
 module as required
 (see  Section  9  and  Ion
 Source PCB Testing  in
 Section  10 of this
 manual).

-------
                                                             Section No.  XIII
                                                             Revision No.   0
                                                             Date:   08/19/5T"
                                                             Page  37 of   37~
                            TABLE XIII-4.  (Continued).
============================================================s===s=========-====

Item         Symptom                Probable Cause           Corrective Action
5.    Manifold tempera-
      ture low
                             Defective manifold heater
                             circuit.
                             Wrong or no temperature
                             data entered in system.
Check manifold heater
connections, heater
and sensor element
continuity, verify
presence of heater
voltage (220 VAC) from
power module.  Repair/
replace defective
components.

Run SS program to
enter temperature
data.
  6.    Manifold tempera-
        ture runaway
                           Wrong or no temperature
                           data in system.

                           Defective temperature
                           cutoff control.
  7.    No heater voltage    Blown fuse.
                             Defective heater triac
                             circuit.
Enter program SS.
Replace temperature
cutoff element.

Replace heater fuse or
triac circuit compo-
nent in power module
(see Section 9 of this
manual).


-------
                                                             Section  No. XIV
                                                             Revision No.~~T"
                                                             Date:  06/01/37"
                                                             Page   1   of   1 "
                                  XI.   SAFETY

Wear
                             ,(
     During any work, the analyst should wear appropriate safety equipment.
Lab coats, safety glasses "and safety shoes should be worn at all  times,
including sample and standard-analysis and preparation.   The operator should
also be aware of hot surfaces (manifold, transfer line,  etc.) and take pre-
cautions against thermal burns.

Gases  ,

     The gases normally used with this instrument are compressed air, helium,
and liquid nitrogen (see Appendix A, Safe Handling of Compressed Gases in the
Laboratory).  The operator and other authorized laboratory personnel  should
exercise caution in handling the liquid nitrogen, as the extremely cold  liquid
can cause severe burns and frostbite.  The helium and air should be handled  as
specified in Appendix A.
Spills
     Spill control pillows should be applied immediately to all spills.   Spills
of extremely toxic substances should be reported to the supervisor immediately
after cleanup  (see Appendix B).

-------
                                                            Section No.  XV
                                                            Revision No.   0
                                                            Date:   08/19/83"
                                                            Page   l of    l~
                                 Miscellaneous

A.   File Deletion In ID0S

     1.   Sometimes It 1s necessary to delete files In ID0S.

     2.   Any files which aren't protected should be protected first.

     3.   > DIAG (CR)
          + PRO LIBRARYNB.LX/33 (CR)

     4.   Still in ID0S you are ready to delete.
          + DE 0:

          This command will give you a file 0.-82VOAXXX.XX
          Type (Y) if you want to delete just type (CR) and it will  now list
          the next file for delection.

     5.   Skip all protected files or this will stop DE and give you a (+)•

     6.   To get out of IDOS:
          + M (CR)

-------
                                       Section No.  A
                                       Revision No.  1
                                       Date:  06/01/3F
                                       Page  1 of  12
           APPENDIX A

SAFE HANDLING OF COMPRESSED GASES
        IN THE LABORATORY

-------
                                       Section No.  A
                                       Revision No.  1
                                       Date:  06/01/3T
                                       Page  1 of  12
           APPENDIX A

SAFE HANDLING OF COMPRESSED GASES
        IN THE LABORATORY

-------
                                                               Section No.  'A'
                                                               Revision No.   1
                                                               Date:   Q6/01/8T"
                                                               Page  2 of  12"
  SATE HANPLTNC OF COMPRSSSIP CASES IS ?H£ LABORATORY

(Taken from AXE8JCAS LABORATORY, September. 1978)
Although tha safety record of eha coaprasaad
gas industry is excellent, the questions
raised by tha users of gaa products, and tha
accidents involving these same users, show
that cany of them havo neither learned nor
applied'tho safety measures that would earn
their "license" for handling coeprassad
gaa.  When handled by personnel who ara prep-
orly  trained and aware of the potential
hazards, coapressod gases ara an safe to...
work with as most of the ordinary chemical
liquids and solids normally handled on a
routine basis in any laboratory.

A compressed gas is defined by tho Dope.
of Transportation (DOT) as "any material
or mixture having in the container an ab-
solute pressure exceeding 40 pai at 70°T
or, regardless of the pressure at 70°F,
having an absolute pressure exceeding
104 pal at 130°T; or any Liquid flacaablo
material having a vapor pressure exceed-
ing 40 psi absolute at 100°T as determined
by AS7M Test D-J23." .
                     i
Hazards
The handling of compressed gases must bo
considered more hazardous than the hand-
ling of liquid and solid materials be-
cause of the following properties
unique  to compressed gases; pressure.,
diffusivity, low flash points for flaa-
mable gases, lev boiling points, and no
visual  and/or  odor detection of many
hazardous gases.  Hazards may arise  aa
a result of equipment failure and leakage
 froa systems  that are not pressure-
 tight.  Also,  improper pressure  control
may cause unsafe  reaction rates  duo  to
 poor flow control.   Diffusion of leak-
 ing gases may  cause  rapid contamination
 of tho  atmosphere, giving riso  to
 toxicity, anesthetic effaces, asphyxi-
ation, and rapid formation of explosive
concentrations of flammable gases.  Th<
flash point of a flammable gas under
pressure is always lower than ambient
or rooa teaperature.  Leaking gas can
therefore rapidly form an explosive
mixture with air.

Lev-boiling-point materials can cause
frostbite on contact with living tissui
This is common among the cryogenic
liquids such as nitrogen and oxygen,
but it also can result froa contact
of the liquid phase of liquefied gases
such as carbon dioxide, fluorocarbons,
and propyleae.  Some compressed gases
are similar to other chemicals in that
they are corrosive,'irritating, and
highly reactive.

The procedures adopted for tho safe
handling of compressed gases are main!
centered on cootaicmont of the aatoria
to prevent its escape to atmosphere,
and proper control  of pressure and
flow.  All rules  and regulations  are
directed toward  these ends.  Knowledge
of emargeacy  procedures is iaportant
to liait property damage or injury, bu
is usually necessary only because a
basic rule of handling has been broken
It is far better to observe tho rules
and avoid  the need  for emergency  aea-
sures.  A  listing of some cosaoa  vio-
lations of basic rules for handling
compressed gases is given in Table 1.

-------
 COKP81SSO CASES (cent.)
                 Section No.  A
                 Revision No.T
                 Date:  06/01/1?
                 Page  3 of  12
Table 1
         Common OSHA violations
         involving comprassod gas
L una«curod crttneon
I CrfcfldersatorodoitnowtpnMoctivocaM
X Noncompatibto gaaea (»wen as ftrdrogon end oaroon)
  stored tog«iner
4. CrKndvrvatosoban who* cylinder tenet in USA (en ah
  tacAcd roowiaior «KtA a tioa«d ditctonjo *e««o is net
1 Rr« citine visners «o« areaon! during voiding, burning.
  or brazing operations
& NO sal try sno««f« and oyowasn tounsams oAoro
  earre>i*o gaa«« are uMd
T. NO ga* masks anojor t««niiy to&w ed n«ef creaa vhoto tox
         «aoo or atorod
Crllndera
Tba oupplior ooako to asoura  tha aafacy
of cylindars through adhara&ca to roju-
latieaa  aae forth by tho Dopartaaat of
Traasportatioa, and by oopplyias cy-
liadoro  with spocifie ralvoa, Labola*
o&d/er oarkinga In oeeerdoaea with
recognised standards.  It  la  mandatory
for tha  ouppllar to ohip cylinders nanu-
factured In conforaance with  DOT
rogulatlooa and to follow  DOT ragulatioe
In tho testing and inspection of cyLinda
tho proper filling of these cylinders,
and tho  uao of safety devices that are
approved by the Bureau of  Explosives.

Figure 1 ahovo cylinder parts and im-
portant  cylinder aarkiags. The cylinder
cap  (1)  protects the cylinder valve.  Tb
valve handwhoel (2) Is uaed  to open and
close tho cylinder valvo.  Valves are
occasionally not equipped  vith hand-
wheels,  and require special vrencheo
to  effect operation.  The  valve packing
sue (3).contains a packing gland and
packing around tho otoa.   It Is adjust  I
only occasionally and Is usually
tightened if leakage la observed around
valve stem.  It should nee be tanpered
with when used In conjunction with
diaphrag»-typo valves.  A safety device
 (4) peralts gas to escape  if the teaper-
aturo gets high enough to  endanger  tho
 cylinder by increased  unsafe pressures.
Tho valve outlet  connection  (5) connect!
 to pressure and/or  flow-regulating
 equipment.  Various  types  of connections
 are provided  to prevent  interchange of
 equlpaenc  for iscoopatiblo gases,
 usually identified  by  CCA (Coap?ess«d
 Cas Association)  nuaber:   for esasple,
 CCA So.  330 Is us«d for hydrogen, «er-
 vieo.   A cylinder cellar (6) holds  the
 cylinder cap  at  all tiaes, except vhea
 regulating eeuiptsent Is attached to
 cylinder v&lv@.   Tha valve outlet csp
 (7) protects valve  threads fro*-

-------
COKP&TSSED CASES (coat)
                                                                Section No.  A
                                                                Revision No.T
                                                                Date:  06/01/82"
                                                                      4 of  12
and keeps the outioc clean; is to nee
used universally.

Specification nuabor (8) signifies chat
the cylinder conforao eo the Department
of Transportation specification POT-3A,
governing svaterlala of construction,
capacities, and tost procedures, ,an4
that tho service pressure for which tho
cyllndar la designed la 2265 psig at ?0°F.

The eylladac serial ttusbar io Indicated
by (9), and (10) Indicates tho dato
(aonth and year: la thio case, Juno
1975) of Initial hydrostatic tenting.
Thar oaf tar, hydrostatic pressure testa
are performed, on cylinders, for moat
gases, every .5 years to deteralne their
fitness for further use.  At this time
nov test dates are stamped Into tho
shoulder of the cylinder.  The original
inspector* a Insignia for conducting
hydrostatic and other required testa to
approve the cylinder under DOT specif 1-
caciona la shovn by (11) •

rilling
Nonliq.uefied  gaaes say bo filled to tho
service pressure marked on a cylinder.
These markings will appear on tho
shoulder of the cylinder, i.e.,
MT3A-2000, indicating  that the cylia-
der has been  manufactured in accordance
with DOT specification  3A, and the
 cylinder filling pressure  la 2000 pal
 at 70°?.

 Since  cospressed gaa  cylinders aro
 handled by a number of  dlf ferene
 types  of plant personnel,  it might
 be veil to consider tho precautions
 to be  takan in b^M^Ttg f roa tho  tiao
 It is delivered until tho tiao it la
 asp tied and ready for return.
Receipt and eontant Identification
When a cylinder la received, it
should have 1) content Identification
by stenciling or labela, 2) a DOT
label, and 3) a valvo-protoction
cap*  Under-no circumstances ahould
tho moana of Identification bo reaovod
froa tho cylinder.  The valve-protection
cap (Tlguro 1) should also regain in
plica until tha user has secured the
cylinder and is ready to withdraw.the
contents.  DOT labels are required
for cylinders in interstate trans*
porcatioa.  Some states require these
labela for intrastate shipments also.
These labels have a mini mm of pre-
cautionary handling inforaation and will
classify the cylinder contents as
flrmahlo, nouflasaablo, poison, or
corroaivo.

Unfortunately, there la aa yet no uai-
fomicy la tho identification of
cylinder contents, although a standard
for «L"*'1p<"S coapressed gases is avail*
ablo.  Soae auppliera provide adequate
atenclling or labela with as such Infor-
mation on thea aa possible, warning
against possible hazards associated
with  tho cylinder contents.  On tho
other hand, cylinders say bo received
with  no identification other than a
color code.   Under no circumstances
should such cylinders be accepted.
Color codes are  of value only la help-
ing the supplier to  segregate large
nvsbors of  cylinders into various
gas services.

Proper Stogsjto

After cylinders  are  received, ghsy
 should be stored la a detached  aad
well-ventilated  or opensided building.

-------
COtCPttSSD CASES (coat)
             Section No.
             Revision No
             Date:  06/01/82"
             Page  5 of  12"
Storage buildings or arena should bo firo
resistant, veil ventilated, located away
from sources of ignition or excassivo heat,
and dry.  Such areas should bo prominently
posted vith tho names of tho gases.being
seorod.  Indoor atorage areas ohould not
bo located near boilers, otoaa or hoe
voter pipes, or any aoureen of ignition.
Outdoor storage araas should have tho
proper drainage and should bo protected
from the direct rays of tho sun in local-
ities where high teaperaeures prevail.
Subsurface storage areas should bo
avoided.  Cylinders should bo protected
against tampering by unauthorized
personnel.

Cylinders should bo chained in place or
put in partitioned cells to prevent thea
from falling over.  Where gases of diff-
erent types are stored at tho saao loca-
tion, cylinders should bo grouped by
types of gas and the groups arranged
to take into account tho gases contained—
for example, fla-aablo gases should not
bo stored near oxidizing gases. Stor-
age in a laboratory should bo confined
to only thooo cylinders in use.  la all
cases, storage areas should comply vith
local, state, and municipal requirements
as vail as with tho standards of tbo
Compressed Gas Association and tho
Bational Fire. Protection Association.

Transportation

When cylinders are bains moved fros a
storage area into tbo laboratory, the
valve-protection cap should bo loft in
place.  Tho cylinder should then bo trans-
ported by moans of a suitable hand truck.
Such a hand truck should bo provided
vith a chain or belt for aocuring tho
cylinder on tho truck.  Do not lift
cylinders by tho cap.  Avoid dragging
or  sliding cylinders.  Use hand trucks
even for short distances.
Securing Cylinder Prior To
Vhoa the cylinder has reached ita des-
tination in the laboratory, it should
bo secured to a vail, a bench, or son*
other fira support, or placed in a
cylinder stand.  Onco tho cylinder ha;
boon secured, tho cap may bo removed,
exposing tha valve.  Tho number of
cylinders 'in a laboratory should bo
liaitod, to ainlatte the firo and
toxicity bazardo.

Safety Relief Devices

Safety devices ere incorporated in
all DOT cot? r eased gas cylinders,
except thoso containing poison or
toxic gao, vhoro tho risk of exposure
to fuaea ia considered more hazardous
than that of a potential cylinder
failure.

Uhoro safety devices are required to
moot DOT regulations , they must bo
approved by the Bureau of Explosives.
Theso saf ocy devices are of four
basic typos:

  1.  Spring-leaded  safety relief,
used mostly for lev-pressure, liquefi
flacaablo gasea.
   2.   Frangible disk, used mostly for
 high-pressure cylinders.

   3.   Frangible disk backed up by a
 fusiblo metal.

   6.   Fusible metal.

 Tho spring-loaded safety  relief  typo
 consists of a spring-loaded seat that
 opens to relievo excessively high
 pressures and then closes vhon tho
 pressure returns to a safo value.
 Tho frangible disk vill burst Jut

-------
COMPRESSES CXSCS (coat)
                 Section No.  A
                 Revision No.~T
                 Date:   gs/Ol/IT"
                 Page 6 of  12
considerably above the oarrico pressure
but not balov or at the hydrostatic coat
pressure of a cylinder, and will release
tho entire cylinder coat onto.  Tho
frangible disk backed op by a fuatblo  .
ootal will fuactioa only If tho tempera-
ture la high enough to malt tho fusible
aatali afear which excessive pceaourao
«U1 burst tha disk, resulting la
release of tha aatlro eoataato of tho
eylladar.  Tha fusible ratal devices
ealt at axeaaalvo temperatures (approx-
imately 160-212°F), allowing tho oatlra
contents of tho cyliadar to escape.

Any of theao devices will prevent a
eylladar from bursting duo to exeessive-
ly high tasperaturea.  Bowovar, tho
latter .two devices may not prevent a
cylinder from bursting la eaaaa whara
aa overfilled cyliadar la exposed to
a temperature that la excessive but not
high enough to malt tho fuaibla metal
aafocy.devices.  Since tho proper func-
tioning of cylinder safety devices depeada
to a largo extant on tho proper f11ling
of a cyliadar, such filling should
aev«z bo attenpted by tho user, unless
express permission hao beea obtained
from the gas supplier.  Safety devices
may also fall to function properly if am
intense flaae irpinging oa the side vail
of a cylinder weakens tha metal to tho
point of failure before heat or pressure
can cause the safety device to function
properly.  Tlnally, it must bo empha-
sized that taspering with cylinder
safety devices la extremely hazardous.

Knowing Tho Caa To  Be Handled
It is of  the ucaaat isportanco that tha
uaer bo vail avaro  of  thoso properties
of a compressed gaa that represent
Hazards  (such aa  flasoability, toxicity,
choaical activity,  and corrosive
effects).  Every  attespt should bo
made to  learn these various  properties
before tho gaa is put  to use./ It la
 sooetiaefl difficult to determine tho
major hazard of aay one gas, siaco this
factor la influenced a great deal by
how tha gaa la used.  In a laboratory
la tho presence of aa opaa flame, tho
f lasaabillty of carbon monoxide might
wall bo tho major hazard, whereas in a
pilot-plant run using carboa monoxide
aa a roactaat, leakage, and therefore
toxicity, may represent tho major hazard

Figure 2 shown tho flaemability ranges
of various gases.  Although the flaa-
mability ranges of the liquefied petro-
leum gases such aa butan* and propane
are relatively short, only very small
concentrations are necessary to ereato
flasaable mixtures.  Tha flaaaability
ranges of acetylene, carbon monoxide,
othylene.oxide, hydrogen sulfide, and
hydrogen are extremely long. Indict-
ing that they can fora explosive
mixtures with air under a wide variety
of conditions.
               l«WlT« O*
OUTAMtt

ttOOVTAMft

OUTlNtt
CAAOOM MOMOUOC
.MTOBOOCN

MTQMOCtM SXA^IBt
              0 •« >• SO •• M *• »«
                  CAS % IW *«
                                      H
      Figure 2

-------
COKPKESSED CASES (cont)
                                                                 Section No.  A
                                                                 Revision No.  1
                                                                 Date:  06/01/8F
                                                                 Page  7 of  1?"
It la important to knev what materials of
construction muat bo unad with a gas to
provoat failure of oquipmoat duo to corro-
aion, or to avoid poaaibla formation of
hazardous eoapousda (ouch ao acetylidas
formed by the reaction ef coppor with
acetylene or gases containing acetylene  .
ta aa ispurity) or tho possible formation
ef fulminates when mercury la used la tho
presence of rrmrmia..            •'
          '«[
Tho hararda of toxic, flairniMo, and
corrosive gaseo can bo minimized by
working la vull-ventilated areaa.  Vhoro
possible, work should bo dona la a hoed,
employing cyllador siiaa that will asauro
uso of *\\ the gaa vlthla a raaooaablo
aaouat of tiao.  Loako ahouLd not bo
ailovad to so oaehoekad.  Advtao tho
attpplisr iasodlately of eyllador leako
that cnnnot bo atoppod by slaplo adjuat-
moat, auca aa tightoning a packlag aot.
        f
 •rooar Plachxrco of Cylinder Contenca
.or controllad reaoval of tho liquid phaao
of a liquofiod gaa, a mnttiial control la
uaod.  Spatial liquid flow regulators
aro alao availablo.  It Boat bo rtusoaborod
that withdrawal of liquid Bust aocasaarily
bo carriod out at  tho vapor pressure* of
tho oatarlal.  Aay attoapt to redueo tho
pruoura will result la flaahiaj ef all
or part of tbo liquid to tho gaa phaao.

 Safaty rcliaf davicta should bo iaatallod
 la all liquid-transfor  liaoa to rollavo
 auddcn,  dangerous hydrostatlc-or vapor*
 prvasuro buildnpo.

 Tor noaliquofiad gasoa, tho aoae eooaoa
 dovico used to roduco prasauzo Co a
 aafo valuo for gaa rnaoval la  a prosauro
 ragulator.  It consists of a spring-
 (or gas*)  loaded diaphragm that controlo
 tho throttling of aa orfico.   Delivery
 pros aura will exactly balance tho dollvdry
 pressure spring to give a relatively
 constant delivery pressure.
Pressure Regulator Handling and Use
A regulator ahould be attached to a
cylinder without forcing the threada.
If tho inlet of a regulator doaa not
fit tho cylinder outlet, no effort
ahould bo made ta try to ferco
fitting.  A poor fit aay indicate
that the regulator la mot intended
for uso oa tho gao choaon.

Tho following procedure ahould bo used
to obtain tho required delivery
presauro:

  1.  After the regulator has boon
attached to tho cylinder valvo outlet,
tarn tho delivery pressure-adjusting
acrow countorclockviao until it turna
freely.

  2.  Opea tho cylinder valve slowly
until tho tank gauge on tho regulator
ragiataro-tho cylinder pressure.  At
thia point, tho cylinder pressuro
ahould be checked ta oeo if it la at
tho expected value.  A largo error may
indicato that tho cylinder valvo la  --\
   3.  With tho flow-control valvo at
 tho regulator outlet closed,  turn the
 delivery pressure-adjusting acrev
 clockviao until the required  delivery
 preasuro la reached.  Control of flow
 g^n be regulated by Banna of  a valvo
 supplied la the regulator outlet or bj
 a ffuppleaentary valve put la  a pipelii
 downstream from the' regulator.  The
 regulator itself should not bo used
 aa a flev control by adjusting tho
 presauro to obtain different  flev
 ratea.  Thia defeato tho purpose of
 tho preasuro regulator, and la soso
 caaaa whore higher flovo are  obtaiaei
 la thia manner, tho presauro  sotting
 may bo in exceso of the:dosign proa-
 sura of tho STStaa.

-------
 COKPS£SSn> CASES  (coot)
              Section No.  A
              JRev1s1on No.~T
              Date:  06/01/oT"
              Page  8 of  12
 Types  ef Pressure Regulators
 Tbo proper  choice of a regulator depends
 on the delivery-pressure rao4o required,
 tho. degree  ef accuracy of dollvery proa-
 ouro to bo  maintained, and tho flow rato
-required.   There ore two basic typoo of
 pressure regulators, single-stage oaH
 double, or  tvo-stago.  The single-
 stage  typo  vill ahov a alight variation
 la delivery pressure as tho cylinder
 pressure dropa.  It vill alao show a
 greater drop in dftlivory pressure than
 a two-stage regulator aa the flow rata
 lo increased,  la addition, it vill ahov
 a higher "lock-up" pressure (pressure
 increase above the delivery oet-point
 necessary to atop flow) than tho eve*
 stage  regulator.  In general, the tvo-
 atage  regulator vill deliver a eoro
 nearly constant pressure under BOTO atria-
 gent operating conditions than vill the
 aingle-atage regulator.

 Peterainatieo ef Cylinder Content
       *
 Aa the content of a cylinder of non-
 liquefied gas ia" discharged, the cylinder
 proaaura decreases by an asount propor-
 tional to the aaount vithdravn.  Tho
 cylinder should be considered eapty
 while poaitivo pressure (25 paig or
 greater) atill roaains, in order to
 prevent auck-back and contamination.
 failure to  close tho valvo on aa
 eapty cylinder vill  allow air and  naia-
 turo to bo dravn into  tho cylinder aa
 it "breathes"  during teaperature
 changes;  an exploaive mixture may
 build up if the gas  is  fli-nablo;  and
 an oxtreaely corrosive condition vill
 be created in cylinders that contain
 chlorine,  hydrogen chloride, or other
 acid-forming or corrosive, gases.

 Aa the vapor phase of a liquefied gaa
 is vithdravn from a cylinder,  tho cyl-
  inder preaauro or vapor pressure vill
 reaain constant as long as any liquid
 is present.  This condition holds true
if tho teaperaturo does not vary.  I£,
however, the material ia vithdravn
froo the cylinder at a rapid rate, the
material itself vill supply tho heat
for vaporization, end upon subsequent
cooling, the vapor proaouro vill bo
lowered.  It io, therefore, iapeaaible
to determine the content of a cylinder
containing a liquefied gas, except by
weighing.  Cylinders containing
liquefied gases are staaped or tagged
vith tho tare weight in order to allow
tho content to be determined.

An indication of cylinder-content
depletion for oomo high preaauro
liquefied gases ouch as carbon dioxide
ethane, and nitrous oxide can bo ob-
tained by noting the cylinder preasurt
After depletion of the liquid phaoo,  '
the cylinder pressure will decrease
below the normal vapor pressure.  A
cylinder containing carbon dioxide vU
have approximately 201 of ito original
content remainingafter depletion of
tho liquid phase.  Aa vith cyLlndero
of nonliqnefied gases, cylinders
containing liquefied gases should
never bo completely eaptied, in order
to prevent truck-back and contaainatiot
of tho cylinder.

Handling tapty Cylinders
Whero cylinders are considered esaty,
the valves should bo closed.  Valve-
protection capa, outlet dust caps, an
other accessories shipped vith the
cylinder should be attached to the
cylinder as received.  Tho cylinder
should be marked or labeled "empty."
Cylinders should then bo placed  in a
proper  storage area, segregated  fron
full cylinders, to await pickup  for
return to the aupplior. :'
     i
Careleaaneao  in the h«**i<;«ig of  an
eapty  cylinder could result in ito
being mistaken for a full cylinder.
Tha connecting of an eapsy cylinder t

-------
 COXPttSSO CASES (coat)
              Section No.
              Revision No."
              Date:   06/01/IT"
              Page  9 of ^12
 a high-prosauro ayatea could eauoo foreign
 maceriala to back up lace tbo cylinder,
 rooultino la oil tha attendant hazarda of
 ouek-back, and tbo pooalblo violent reaction
 within the cylinder.

 La ak Potaction
 Chock cylinders and all eonnoctiona uador
 proaaura for leaks prior to uaing cho
 contents.  Whoa using toxic gasea, it in
 adrisablo that aoaa dovico or indicator
 ba uood to warn of tha presence of toxic
 coneantrations.  For example, otrlpa of
 lead acetate paper can bo hung in an
 aroa vharo hydrogen aulfido ia being
 uaod.  Although tola gaa haa a disagreeable
Oodor, it ooon deadena tho aonso of small,
 rendering tho uaar incapable of detecting
 increasing dangarbua concentrations by
 odor.

 Cyliadara that develop looks ahould bo
 treated aa follova:  Cylinder valve—
 packing leaks of acidic gasea, ouch an
 chlorine, hydrogen chloride, hydrogen
 aulfido, and aulfur dioxide, can uaually
 bo corractod by tightening tha valvo
 packing nut (turn clockviso aa vievod
 froa abovo).  If Talvo laaka poralat or if
 loaki appoar at any portion of tho
 cylinder, adyiae the aupplior isaodiatoly.
 E*move tha cylinder to a hood or location
 where the leakage cannot eautto daaago
 until tho coneonta can aafety bo dia-
 peaod of by vonting or abaorption in wntor
 or caustic oolution or until information
 on other maana of dispooal aa rococanoded
 by the  supplier ia obtained.  Cylinder
 valve-packing leaka of basic gaaoo
  (asaonia  and xainea) can bo atopped by
 tightening tho packing out  (turn counter*
 clockviao aa viewod  fron above).  Laako
 of those  basic gasea through tha vulva
 outlet  that cannot bo controlled by
  turning the handwhaol to tho clooed
  position ahould be  stopped by plugging
  the  outlet with  3/8-in pipe plugs.  12
  laaka persist, roaavo the cylinder to a
•hood and ask cho auppliar for inform-
ation en bow to resedy tha problea.

On raro occaaiona, oaargoacy action
aay bo aocoasary ia order to aove a
leaking cylinder, to a location where
it can vent oafely, or it oay have to
bo rcsoved f roo a building and brought
eutdoora.  tn ouch inatancaa, aa
eaorgency plan should bo put into
offoct:

1.  Properly warn all personnel
required to evacuate a building or
aection of a building*

2.  Shut off electrical power to
prevent ignition of a leaking flaa-
mablo
3.  Dotaraine the abort eat  route so
the point of gaa dlapoaal.

A.  Obtain satisfactory  conveyance,
ouch  an a hand  truck,  to novo tha
cylinder aviftly.

5.  Peat the area whora  the cylinder
ia venting to prevent  taaparing by
unauthorised personnel.

Such  an oaargency plan can  function
efficiently only if a  trained safety
crew  ia educated in the  proper hand-
ling  of gaa cylinders, with training
ia tho procedures to bo  followed in
caaea of osargancy  with  all tho g*saa
handled by tho  facility. £a.uipaent
ouch  aa self-contained gaa  masks
must  bo available for  *^**"<11ng toxic
 gasea or for handling asphyxiating
 gaaoa in cloao confinati.   Esarganciei
 involving fln-nnbla gasea aunt  bo
 managed with tho ucaoat care in order
 to prevent ignition.  The afteraath
 ef sroaa leakage ia axtreaely iaport
 AH *ra*a aoat bo adaquately vented
 before the restoration of pover ia

-------
CASK (cant)
                                                              Section No.  A
                                                              Revision No.  1
                                                              Date:  06/01/3F
                                                              Page 10 of  12
coses of flasmable-gas leakage.  Areas
contaainated by corrosive gases oust bo
adequately vented and completely washed
down-to prevent subsequent degradation
of delicate instruments, electrical
contacts, ace.

Handling of Corrosive Cases
Corrosive gases should bo scored for tho
shortest possible periods before use,
preferably less then throe months.
Storage areas should bo as dry as possible.
A good supply of water should bo available
to handle oaergency leaks.  Moot corro-
sive gases «r«a bo absorbed ia water.

Corrosive gases should not bo stored ia
areas containing instruments or other
devices sensitive to corrosion.  Those
gases should be segregated as CO typo,
and rotated so that tho oldest stock is
used first.  The saalleat cylinders
possible should be used to ensure rea-
sonable turnover of cylinders.  Cylinders
used and then> put back in storage should
have all appurtenances reaoved froa tho
valve outlet and should preferably bo
flushed with dry nitrogen or air to
keep thea ia good working order.

Uhen corrosive gases are being used, Che
cylinder valve stem should bo worked
frequently Co prevenc "freezing."  The
valve should bo closed when tho cylinder
is not in use.  Regulators and valves
should be closed when the cylinder is
not  in use and flushed with dry air or
nitrogen after use.  Such control
devices should not bo left on a cylinder,
except when it is in 'frequent use.
Vhan corrosive gases are  to bo dis-
charged into  a liquid, a  trap, check
valve, or vacuum break device  should
always be employed to prevenc danger-
ous  suck-back.
                                     Occasionally a valve otea does boeoao
                                     frosen and cannot be turned.  Thin
                                     condition aay bo corrected by first
                                     plugging the cylinder valve outlet
                                     with a solid plug, gaskatod cap, or a
                                     closed needle valve, loosening the
                                     volvo packing nut, t id than trying
                                     eo cum the valv_o stun nanunlly with
                                     a corroccly fitted wrench (no longer
                                     than 8 inches), or by slightly tapping
                                     cho wrench with a light cool.  Once
                                     the scea is turning freely, shuc the
                                     valve, tighten the packing nut, and
                                     safety vent the gas crapped in tho
                                     valve pore.  This operation should be
                                     dona ia a hood or out of doors.  If
                                     tha cylinder valve cannot bo unfrozen
                                     by ehis procedure, ask the supplier foi
                                     indorsation or return cho cylinder.

                                     General precautions for handling,
                                     storing, and using compress>td gases
                                     is presented separately ia Attachaenc
                                     One.  You may post it ia your work
                                     area as a daily reaindar.

-------
        PZECXDTIONS Dt BAMfiUSC
OF COiffSlSSED CASZS ZH IBS
LABOKAIOBX (coat)
                                                               Section No.  A
                                                               Revision No.~T
                                                               Date:  06/01/52"
                                                               Page ll~oT IF
          use when working vith toxic
gases.  Those masks obould bo placed
la convenient locations ia areas net
likely to become contaminated, and
obould bo approved by tho U.S. Bureau
of Mines for, tho oorrlco intendod.
Those involved.in the handling of
compressed gases should bocomo
familiar with the proper application
aad Uaitatioaa of the various typoa
of aaalta aad raspiration aido avail-
oblo.

13.  Vhoa. raturciag cylinders, clooo
tho valve before abipaene, loaviag
oome poaitive prasouro in tbo cy-
liader.  K«place any valve outlot
and protective capo originally
shipped vi'th tao cyliader.  Mark
or label the cyliader "ompty"
(or utilize 'standard DOT "eapty"
labels) aad store in a designated
area for return to tho supplier.

16.  Before uuiag cyliadora, read
all label information and data
sheets associated vith the gas
being used.  Observe all applicable
safety practices.

17.   Eye baths, safety shavers,
gas auks, respirators, and/or
rususcitators should be located
nearby but out of  tho iracdiato
aroa that is  likaly to bocoso
coataainatod  in the ev«nc  of a
largo release of  gao.

IS.   Fire oxtinguishera, prefer-
ably of tho  dry  choaical  typo,
should be kept clooo at brrad and
 should be chocked periodically
 to assure eh«is  proper  operation.
Tho user of conpressod gases should
become faailiar with tho first-aid
cathodo to bo employed in cases of
ovorezpoouro or bvmo caused by a
Sao.  A plant doctor obould bo
familiar vith whatever further creat-
aaato may bo necessary.  Unnecessary
delay ia the treataent of a patieat
overcome by a toxic gas or burned
by a corrosive gas could cause tho
patieat peraaaeat daaage aad night
oven result ia death.  Authorized
personnel should odainister first aid;
hovovor, they should not take it upoe
themselves to adsinister medical
treatments.  A physician should bo
contacted immediately.

-------
                                                        gyr ONE
CCfTBAL PRECAUTIONS IS BA2DLZKC
Of COXPSESSO CASES DC THE
UkBOSAXORt
              Section No.  A
              Revision No. __!__
              Date:  06/01/82
              Page 12 of  12
Scmo general precautions for haadliag,
storing, aad using cospreaaod gasaa
follow.

1.  B«vor drop cylinders or paralc thes
eo strike oasa other violently.

2.  Cylinders aay bo stored la tbo opaa.
but should bo procoecad from eha ground
beneath to prevent rusting.  Cylinders
My bo stored la eha nun, except la
localities vharo extreaa tasparaeuraa
prevail; la tha auto of curtain gaaeo,
eha supplier'a rocesaeadatiba for
shading should bo obaerved.  If lea or
snov accusulates on a cyllndar, chav
ac rooa cesperacura or with vaear at a
te^aratuxa not axcaedlag 125°f.

3.  Tha valve-protoctioa cap ahould bo
loft on oach cylinder until le haa beaa
securod agnfnte a vail or boneh, or
placed la a cylinder otaad, and la
raady to bo used.

4.  Avoid dragging, rolling, or oilding
cylinders, oven for a short distance.
They should be novod by using a suitable
hand track.

S.  Kevar tsspar vith safety devlcos la
valves  or cylinders.

6.  Do  not  store  full and  oapty cylinders
together*   Serious  suck-back can occur
when aa espty cylinder  la  attached to
a prassurixed oyotc
 7.  So part of a cylinder ahould bo sub-
 jected to a tosperaturo higher tbaa 122°?.
 A flasa should never bo permitted to coma
 la contact vita any part of a cosprassod
 gao cylinder.
 8.   Cylinders should act be subjected
 to  artificially created lev teapera-
 turoo (-20 T or lover), olaco aaay
 typos of otaol vlll lose their
 ductility aad ispact strength at low
 tosporatureo.  Special atainlais steel
 eyliodara'arft available for lev tea-
 peraturo uao.

 9.   Do not place cylinders vhere they
 aay bacoao part of aa electric circuit
 When electric are-veldiag, precautions
 oust bo taken to prevent striking aa
 arc ogainst a cyliadar.

10..  Bead aad ground Ml cylinders,
lines, aad oquipaeat used vlth
flasaabla coepressad gases.

11.   Uso coopresaed gases only la voll-
voatilated area.  Toxic, flaasabla, ygd
corrosive gases should bo handled la
a hood.  Only snail cylinders of toxic
gooes ahould bo used.

12.  Cylinders should bo used la rotatit
as received froa the supplier.  Storage
areas should bo see up to perslt invont-
cry rotaeloa.

13.  Vhea discharging gas into a liquid
a trap or suitable check valve should b<
used to prevent liquid froa  getting
back into the cylinder or regulator.

14.  Vhea using compressed gases, vear
appropriate protective equipment, such
as mfaE? fogglss or face shield,
rubber gloves, aad safety shoes,  tfall-
veasUUtsd barricades should bo us ad la
excreaaly hazardous operations, such
aa ia the handling of fluorine.  Cad
easka ^'-Uti b« kept available for

-------
                           Section No.  8
                           Revision No."~T
                           Date:  06/01/5F
                           Page  1 of  2
APPENDIX B

-------