Federal Water Pollution Control  Administration
             Division of Research
       Analytical Quality Control Branch
   FWPCA OFFICIAL INTERIM  METHODS
For Chemical Analysis  of Surface Waters
            September 1968

-------
             Federal Water Pollution Control Administration



                   Analytical Quality Control Branch



                       OFFICIAL INTERIM METHODS



                                  for



                   Analyses of Surface Water Samples





                                         Storet Parameter No.   QQ435



Parameter    :   Acidity,  Total (mg/1 as CaC03)
Method
•  Volumetric, with electrometric endpoint.
Reference    :  l.   Standard Methods for the Examination of Water and

                    Wastewater,  12th ed.,  APHA,  Inc., N.Y., 1965,

                    46-47.


                2.   ASTM 1968 preprint,  D1067,  pp.  38-41.
Modifications:   None
                                                _  _    September 1968
                                                Date:    v

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet Parameter No.  00410

Parameter    :  Alkalinity, Total (mg/1 as CaC03)
Method
:   Volumetric, with electrometric endpoint.
Reference
   1.  Standard Methods for the Examination of Water and
      Wastewater, 12th ed., APHA, Inc., N.Y., 1965,
      48-52.

   2.  ASTM  1968 preprint, D1067, pp. 38-41.
Modifications:  None
                                                Date:
                                                      September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No. to be  assigned
Parameter    :   Aluminum,  Dissolved (yg/1)
Method
Reference
                Filtration through 0.45  y  membrane filter.
                Measurement by  atomic absorption.
                Method is specific to the particular instrument.
                General procedure is  being selected.
Modifications:  Use of 0>45 y membrane filter only.
                                                Date:  September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No.   01105
Parameter    :  Aluminum,  Total Cyg/D
Method
Atomic absorption.
Reference    :  Method is specific to the particular  instrument,
                General procedure is  being selected.
Modifications:  None
                                                Date;  September  1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No. to be  assigned
Parameter    :  Arsenic,  Dissolved Cug/D
Method       :  Filtration through 0.45  y  membrane filter.
                Measurement by silver diethyldithiocarbamate
                  colorimetric method.

Reference    :  standard Methods for the Examination of Water and
                Wastewater, 12th ed., APHA,  Inc.,  N.Y., 1965,
                56-57.
Modifications:  Filtration through 0.45 p membrane filter only.
                                                Date:  September 1968

-------
             Federal Water  Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet  Parameter No. QlOQQ
Parameter    :   Arsenic, Total,  (yg/1)
Method
Measurement by silver diethyldithiocarbamate
colorimetric method.
Reference    :   Standard Methods  for  the Examination  of Water  and
                Wastewater,  12th  ed., APHA, Inc., N.Y., 1965,
                56-57.
Modifications:  None
                                                Date:  September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet Parameter No.  00525

Parameter    :   Ash,  Solids Dissolved (mg/1)
Method
Gravimetric, ignition at 600 C, modified.
Reference    :   Standard Methods for the Examination of Water and
                Wastewater,  12th ed., APHA,  Inc.,  N.Y., 1965,
                247-248.
Modifications:  Use of 0.45 y membrane filter only, for solids
                separation.
                                               Date;  September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet Parameter No.  00540

Parameter    :  Ash, Solids, Suspended (mg/1)



Method       :  Gravimetric, ignition at 600 C, modified.
Reference    :  Standard Methods for the Examination of Water and
                Wastewater, 12th ed., APHA, Inc., N.Y., 1965,
                247-248.
Modifications:  use of 0.45 y membrane fLiter only for solids
                separation.
                                               Date;  September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet Parameter No.   00510

Parameter    :   Ash,  Solids,  Total (mg/1)



Method       :   Gravimetric,  ignition at 600 C.
Reference    :   Standard Methods  for the Examination of Water and
                Wastewater,  12th  ed.,  APHA,  Inc.,  N.Y.,  1965,
                247-248.
Modifications:   None
                                                Date;  September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet Parameter No.  00310

Parameter    :   Biochemical Oxygen Demand, 5 day (mg/1)
Method
Non-official method - modified Winkler - azide method,
Reference    :   Standard Methods for the Examination of Water and Waste-
                water, 12th ed., APHA, Inc., N.Y., 1965, 406-410; 415-421.
Modifications:  1.  Full Bottle Technique
                2.  Scope and Application;  For the BOD test to be most
                meaningful the laboratory procedure should duplicate
                environmental conditions -- temperature, mineral sub-
                strate, water movement, biological population, etc.  If
                determination of the BOD in this manner was carried out
                the procedure would be considered "standard"; however,
                since it would be extremely difficult to reproduce these
                conditions from one laboratory to another the test must
                be recognized as a non-standard procedure, and there-
                fore classified as "Non-Official".

                3.  Apparatus:  Incubation bottles, 300 ml + 3 ml
                capacity, with ground-glass stoppers.

                4.  Data Validity:  The most valid BOD results are
                obtained when samples have at least 2 mg/1 of depletion
                                                Date;  September 1968

-------
and at least 1 mg/1 of oxygen remaining.  Values
that do not meet these criteria should be reported
with the Subscript L - meaning that either the
depletion or residual, were below the accepted
limits.

     The minimum number of dilutions that should
be run on any sample is two - the preferable number
is three.  In all cases the only figures that
should be used or averaged are those falling within
the desired or minimum ranges.  If all resulting
values are below the minimum, then these data will
be averaged and reported with the proper connotation.

     Example: 3 dilutions - 2 within limits.
               1 below
              Average, only those two falling
               within limits.

              3 dilutions - 1 within. 2 below
              Report only the value falling
               within the limits.

              3 dilutions - 3 below
              Average all three, report value
               as BOD.

              2 dilutions - 1 within. 1 below
              Report only the value falling
               within the limits.

              2 dilutions - 2 within
              Average both values.

5.  Precision and Accuracy:  The expected standard
deviation of the "Standard Methods" BOD procedure
is approximately + 20 percent of the mean value .
At present, data are not available to substantiate
whether or not the full bottle technique will alter
this reported deviation.  However, it is felt that
since  the FWPCA procedure does not significantly
alter  the mechanics or chemistry of the Standard
Methods procedure, the precision and accuracy will
not be seriously affected.
 Ballinger, D.G., Lishka, R.J., Reliability and
 Precision of BOD and COD Determinations;
 Journal WPCF, May 1962, pp 470-474.            $[p    10-

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No. _ta_hfi_assigned
Parameter    :  Cadmium, Dissolved (yg/1)
Method       :  Filtration through. 0.45 y membrane filter.
               Measurement by atomic absorption.
Reference    :  Method is specific to the particular instrument.
               General procedure is being selected.
Modifications: Use of 0.45 y membrane filter only.
                                                Date:  September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No. °1025
Parameter    :   Cadmium,  Total Cyg/D
Method
             :   Atomic absorption.
Reference    :   Method is specific to the particular instrument.
                General procedure is  being selected.
Modifications:  None
                                                Date.  September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet Parameter No.  to  be assigned

Parameter    :  Calcium,  Dissolved,  as Ca,  (mg/1)
Method       :  Filtration through 0.45 y  membrane filter.
                Measurement by atomic absorption.
Reference    :  Method is specific to the particular instrument.
                General procedure is being selected.
Modifications:  Use of 0.45 y membrane filter only.
                                                Date:  September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet Parameter No.  00915

Parameter    :  Calcium, Total,  as Ca,  (mg/1)
Method
             : Atomic  absorption.
Reference
             :  Method  is specific  to  the particular instrument.
               General procedure is being selected.
Modifications:  None
                                                Date: September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet Parameter No. QQ340

Parameter    :  Chemical  Oxygen Demand,  High Level  (mg/1)
                Official  Interim Procedure  #1
Method
:  Volumetric, dichromate reflux.
Reference    :  1.   Standard Methods  for  the Examination  of Water  and
                    Wastewater,  12th  ed., APHA,  Inc., N.Y., 1965,
                    510-514.

                2.   ASTM Book  of Standards, Part 23,  1967
                    D1252-67,  pp.  244-247.

Modifications:  None
                                                Date:  September 1968

-------
Reference
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet Parameter No.  00335
Parameter    :  Chemical Oxygen Demand,  Low Level (mg/1)
                Official Interim Procedure #2
Method       :  Volumetric,  dichromate reflux,  modified.
                i.  Standard Methods for the Examination of Water and
                    Wastewater, 12th ed. ,  APHA,  Inc.,  N.Y., 1965,
                    510-514.

                2.  ASTM Book of Standards,  Part 23,  1967
                    D1252-67, pp. 244-247.
Modifications:  See new method write-up.
                                                Date:  SePtember 1968

-------
             Federal Water Pollution Control Administration




                   Analytical Quality Control Branch




                       OFFICIAL INTERIM METHOD




                                  for




                  Chemical Oxygen Demand - Low Level




Interim Official Procedure #2




1.  Scope and Application;




     This method is applicable when the chloride level is less than




100 mg/1 and the COD is in the range of 10-90 mg/1.  Otherwise the




scope and application is exactly as the Interim Official Chemical




Oxygen Demand Procedure.  Extreme care must be taken with this procedure




because a trace of organic matter from the glassware or atmospheric




contamination may cause a gross error.  AH glassware used in the test




should be "aged" in that previous low level blank determinations must have




been run to eliminate traces of organic material.




2.  Apparatus;




     2.1)  Reflux apparatus:  125 ml Erlenmeyer flasks with standard




ground glass 24/40 neck and 300 mm jacket Liebig, West, or equivalent




condensers with 24/40 ground glass joint, and a hot plate or mantle




with sufficient power to produce at least 9 watts/Sq. in. of heating




surface, or equivalent, to insure an adequate boiling of contents of




th£ refluxing flask.




3.  Sampling;  As per Official Interim Procedure #1.




4.  Preservation;  As per Official Interim Procedure #1.




5.  Reagents;




     5.1)  Standard potassium dichromate solution:  0.025N.  Dissolve




12.259 g K2Cr207, primary standard grade, previously dried at 103°C for




2 hours, in distilled water and dilute to 1000 ml.  Mix this solution




                                                               SEP     1968

-------
                               -2-

thoroughly then dilute 100 ml to one liter with distilled water.  See
referee procedure for elimination of nitrite interference.

     5.2)  Sulfuric acid reagent:  Cone.  H2S04 containing 23.5 g silver
sulfate, Ag2S04, per 9 Ib. bottle (1 to 2 days required for dissolution.)

     5.3)  Standard ferrous ammonium sulfate titrant:  0.025N
           5.3.1)  Dissolve 98 g of Fe(NH4)2 • 6H20 in distilled water.

           Add 20 ml of cone. H2S04, cool, and dilute to 1000 ml.  Dilute
           100 ml of this solution to 1 liter with distilled water.  This
           solution must be standardized daily against K^Cr^Oy solution.
           5.3.2)  Standardization:  To 15 ml of distilled water add 10 ml

           of 0.025N Kr2Cr207 solution.  Add 20 ml of H2S04 and cool.

           Titrate with ferrous ammonium sulfate using 1 drop of ferroin

           indicator.  The color change is sharp, going from blue-green
           to reddish-brown.

                         Normality = (ml K2Cr207) (0.025)
                                      ml Fe (NH4)2(S04)2
     5.4)  Ferroin indicator solution:   As per Official Procedure #1.

     5.5)  Silver Sulfate, reagent powder:  As per Official Procedure #1.
     5.6)  Mercuric Sulfate:   As per Official Procedure #1.
     5.7)  Sulfuric Acid:  36N, Reagent grade.
6.    Procedure;

     6.1)  Determine chloride content of sample.  Pipet 10 ml of sample
into the reflux flask.  Add 10 ml of 0.025N Kr2Cr207 solution then 2 ml

of cone. H2SO^ and mix.  Add HgSO^ in the ratio of 10 mg HgS04 to 1 mg
chloride, and swirl until all mercuric sulfate has dissolved.  Add 2
glass beads which have previously been ignited at 600°C for 1 hour.

                                                              SEP     1968

-------
                                -3-

Then add 18 ml of sulfuric acid-silver sulfate solution and gently

swirl until the solution is thoroughly mixed.

     6.2)  Attach the flask to the condenser and reflux for 2 hours.

Cool and wash down the condenser with 3 ml of distilled water.  Discon-

nect the condenser and wash the flask with 2 ml of water.  Cool to room

temperature.

     6.3)  Titrate with standard ferrous ammonium sulfate using 1 drop

of ferroin indicator.

     6.4)  Run a blank using 10 ml of COD-free distilled water through

the entire procedure.

     6.5)  Concentration of dilute samples.  If an increase in sensitivity

is required, it may be obtained by reducing a 20 ml sample volume by

boiling in the refluxing flask on a hot plate in the presence of all

reagents.  A blank should be carried through the same procedure.  This

technique has the advantage of concentrating without significant loss of

easily digested volatile materials.  Hard to digest volatile materials,

such as volatile acids,  are lost,  but an improvement is gained over

ordinary evaporative concentration methods.

7.  Calculation;

                      mg/1 COD = (a - b) c x 8000
                                    ml sample
               Where:  a = ml FeCNH^^ (80^)2 used for blank;

                       b = ml Fe(NH,)2 (S0,)2 used for sample;

                       c = normality of Fe(NH )  (SO.) .

8.  Precision and Accuracy:

     Data are unavailable on the precision and accuracy of this method

at the present time.
                                                              SEP    1963

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No.  to be assigned
Parameter    :   Chemical Oxygen  Demand,  Saline Waters  (mg/1)
                Official Interim Procedure  #3
Method
:  Volumetric, dichromate reflux, modified.
Reference    :   1.   Standard Methods for the Examination of  Water and
                    Wastewater,  12th ed.,  APHA,  Inc.,  N.Y.,  1965,
                    510-514.

                2.   ASTM Book of Standards,  Part 23,  1967
                    D1252-67, pp.  244-247.


Modifications:   See new method write-up.
                                                Date;   September 1968

-------
          Federal Water Pollution Control Administration

                Analytical Quality Control Branch

                    OFFICIAL INTERIM METHOD

                               for

           Chemical Oxygen Demand, Saline Waters (mg/1)
    Official Interim Procedure #3.
1.   Scope and Application;

     When the chloride level exceeds 1000 mg/1 the minimum accepted

value for the COD will be 250 mg/1.  COD levels which fall below this

value are highly questionable because of the high chloride correction

which must be made.  Otherwise the scope and application is exactly as

the Interim Official Chemical Oxygen Demand Procedure #1.

2.   Apparatus;  As per Official Interim Procedure #l.

3.   Sampling;  As per Official Interim Procedure #1.

4.   Preservation;  As per Official Interim Procedure #1.

5.   Reagents;

     5.1)  Standard potassium dichromate solution:  As per Interim

Official Procedure #1.

     5.2)  Sulfuric acid reagent:  Cone. t^SO^ containing 23.5 g silver

sulfate, Ag2SO/, per 9 Ib bottle  (1 to 2 days required for dissolution).

     5.3)  Standard ferrous ammonium sulfate titrant, 0.25N:

           5.3.1)  Dissolve 98 g of Fe(NH4)2(S04)2  • 6H20 in distilled

                   water.  Add 20 ml of cone. l^SO^, cool, and dilute to

                   1000 ml.  This solution must be standardized against

                   the standard potassium dichromate solution daily.
                                                               SEP     1S68

-------
                                -2-

           5.3.2)  Standardization:  Dilute 25.0 ml of standard

                   dichromate solution to about 250 ml with distilled

                   water.   Add 75 ml cone,  sulfuric acid.   Cool then

                   titrate with ferrous ammonium sulfate titrant,  using

                   10 drops of ferroin indicator.

                            Normality = (ml K2Cr20?)(0.25)

                                          ml Fe(NH4)2(S04)2


     5.4)  Ferroin indicator solution:  As per Official Interim Procedure #1,
             •
     5.5)  Silver Sulfate, reagent powder:   As per Official Interim

Procedure #1.

     5.6)  Mercuric Sulfate:  As per Official Interim Procedure #1.

     5.7)  Sulfamic Acid:   As per Official Interim Procedure #1.

     5.8)  Sulfuric Acid:   36N, reagent grade.

6.   Procedure;

     6.1)  Pipet a 50 ml aliquot of sample not to exceed 800 mg/1 of COD

into a 500 ml, flat bottom, g.s. erlenmeyer.  Add 25 ml of 0.25N Kr2Cr207

then 5.0 ml of cone. H2S04 (containing no silver sulfate).   Add HgSOA in

the ratio of 10 mg to 1 mg chloride, based upon the mg of chloride in the

sample aliquot.  Swirl until all the mercuric sulfate has dissolved.

Carefully add 70 ml of sulfuric acid-silver sulfate solution and gently

swirl until the solution is thoroughly mixed.  Glass beads should be

added to the reflux mixture to prevent bumping, which can be severe and

dangerous.  (CAUTION:  The reflux mixture must be thoroughly mixed before

heat is applied.  If this  is not done, local heating occurs in the bottom

of the flask,  and the mixture may be blown out of the condenser).



                                                              SEP    ma

-------
                                 -3-




     6.2)  Attach the flask to the condenser and reflux the mixture for




two hours.




     6.3)  Cool, and wash down the interior of the condenser with 25 ml




of distilled water.  Break the joint of the condenser and wash the flask




and condenser joint with 25 ml of distilled water.  Remove the condenser




and carefully add to the flask 175 ml of distilled water so that the




total volume is 350 ml.  Cool to room temperature.




     6.4)  Titrate with standard ferrous ammonium sulfate using 10 drops




of ferroin indicator.  (This amount must not vary from blank, sample and




standardization).  The color change is sharp, going from blue-green to




reddish brown and should be taken as the end point although the blue-




green color may reappear within minutes.




     6.5)  Run a blank, using 50 ml of distilled water in place of the




sample together with all reagents and subsequent treatment.




     6.6)  For COD values greater than 800 mg/1 a smaller aliquot of




sample should be taken, however, the volume should be readjusted to




50 ml with distilled water having a chloride concentration equal to the




sample.




     6.7)  Chloride correction*- Prepare a standard curve of COD versus




mg/1 of chloride, using sodium chloride solutions of varying concentra-




tions following exactly the procedure outlined.  The chloride interval,




as a minimum should be 4000 mg/1 up to 20,000 mg/1 chloride.   Lesser




intervals or greater concentrations must be run as per the requirements




of the data, but in no case must extrapolation be allowed.








*Burns, E.  R.,  Craig, M.,  Journal WPCF,  Vol. 37, pp 1716-1721, 1967






                                                              SEP     1363

-------
                                 -4-

7.  Calculation;

                  mg/1 COD = (a - b) c x 8000 - d x 1.20
                                  ml sample

         where

                  COD = chemical oxygen demand from dichromate

                   a  = ml Fe(NH^>2 (804)2 for blank;

                   b  = ml FeCNH^^ (SO^^ for sample;

                   c  = normality of Fe^H^^ (S0^)2;

                   d  = chloride correction from curve (step 6.7)

                 1.20 = Compensation factor to account for the extent

                        of chloride oxidation which is dissimilar in

                        systems containing organic and nonorganic

                        material.

8.  Precision and Accuracy;

     Data are unavailable on the precision and accuracy of this method

at the present time.
                                                           SEP     1868

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No.  00940
Parameter    :   Chloride  (mg/1)
Method
:  Volumetric, mercuric nitrate.
Reference    :   1.   Standard Methods  for  the Examination  of Water and
                    Wastewater,  12th  ed., APHA,  Inc., N.Y., 1965,
                    87-90.

                2.   ASTM Book  of Standards, Part 23,  1967
                    D512-67, pp.  25-27.


Modifications:   None
                                               Date;  September  1968

-------
Parameter
Federal Water Pollution Control Administration

      Analytical Quality Control Branch

          OFFICIAL INTERIM METHODS

                     for

      Analyses of Surface Water Samples


                            Storet Parameter No.  00365

:   Chlorine Demand,  15  minutes  (mg/1)
Method       :  No official method selected.   Unofficial method
                is for control of disinfection.
Reference    :  Standard Methods for the Examination of Water and
                Wastewater,  12th ed.,  APHA,  Inc.,  N.Y.,1965,
                381-383.
Modifications:  None
                                                Date;  September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet Parameter No.  00370

Parameter    :  Chlorine Demand, 1 hour  (mg/1)
Method       :  NO official method selected.  Unofficial method
               is for control of disinfection.
Reference    :  Standard Methods for the Examination of Water and
               Wastewater, 12th ed., APHA, Inc., N.Y., 1965,
               381-383.
Modifications: None
                                                Date:  September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet Parameter No.   QQ375

Parameter    :   chlorine Demand,  2  hour (mg/1)
Method       :   No officiai method selected.   Unofficial method
                is for control of disinfection.
Reference    :  standard Methods for the Examination of Water and
                Wastewater,  12th. ed.,  APHA,  Inc.,  N.Y., 1965,
                381-383.
Modifications:   uone
                                                Date:  September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet  Parameter No.  QQ38Q

Parameter    :  chlorine Demand, 24 hour (mg/1)
Method
Reference
               No official method selected.  Unofficial method
               is for control of disinfection.
               Standard Methods for the Examination of Water and
               Wastewater, 12th ed., APHA, Inc., N.Y., 1965,
               381-383.
Modifications: None
                                                Date;  September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet Parameter No.   50060

Parameter    :   Chlorine,  Residual (mg/1)



Method       :   Volumetric, (iodometric)



Reference    :   Standard Methods for the Examination of Water  and
                Wastewater, 12th ed.,  APHA,  Inc.,  N.Y., 1965,
                376-378.

                ASTM Book  of Standards, Part 23 (1967), 249-252.



Modifications:   None
                                                Date.   September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No. to be  assigned
Parameter    :  Chromium,  Dissolved (yg/1)
Method       :  Filtration through Q.45 y  membrane filter.
                Measurement by atomic  absorption.
Reference    :   Method is specific to  the particular instrument.
                General procedure is being selected.
Modifications:  Use of 0.45 y membrane filter only.
                                               Date:  September  1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet Parameter No. Qifnn

Parameter    :   chromium,  Total (yg/1)



Method       :   Atomlc absorption.
Reference    :  Method is specific to the particular instrument.
                General procedure is  being selected.
Modifications:   None
                                                Date:  September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No.  00080
Parameter    :   Color (Pt  -  Co units)
Method
•  Platinum—cobalt.
Reference    :   Standard Methods for the Examination of  Water and
                Wastewater,  12th ed.,  APHA,  Inc.,  N.Y.,  1965,
                127-129.
Modifications:
  Permanent color standards are acceptable if these are
  individually and periodically recalibrated against
  platinum cobalt standards.

  See method comments attached.
                                                Date;  September 1968

-------
            Federal Water Pollution Control Administration
                  Analytical Quality Control Branch
                     OFFICIAL INTERIM METHOD
                                 for
                                Color
Method Selection
     The Platinum-Cobalt Method is selected as the Interim Official
FWPCA Method.  It is the simplest, most rapid, and is the most useful
under all conditions.   The Tristimulus and Spectrophotometric Methods
are useful in detecting specific color problems.
Definitions
2.  The term "color" in these methods is considered to be the color
of the light transmitted by the water or waste water after removing
the suspended material including the pseudocolloidal particles
("true color").
     It is recognized  that the color characteristics of water are
affected by reflectance ("apparent color").  However, until a suitable
method is available for making solution reflectance determinations,
the color measurement  will be limited to the characteristics of light
transmitted by clarified samples.
     The color of the  filtered water is expressed in terms which
describe the sensation realized when viewing the clarified water.
The hue (red, yellow,  green, etc.) of the color is designated by the
term dominant wave length, the degree of brightness by luminance,
and the saturation (pastel, pale,  etc.) by purity.
                                                                    SEP     1968

-------
                                                                  2.
     The light transmission data are converted to the color class-




ification terms by using standards adopted by the International




Commission on Illumination ("CIE"), and the selected-ordinate method.






Purity of Reagents




     3.  Reagent grade chemicals shall be used in all tests.




Sampling




     4.  Samples for the color determination should be representative




and must be taken in clean glassware.  The color determination should




be made within a reasonable period, as biologic changes occurring in




storage may affect the color.

-------
                                                                   3.





                     Platinum-Cobalt Method




Application




     5.  This method is applicable to the determination of color of




water and waste water for routine control purposes.




Principle of Method




     6.  Color is measured by visual comparison of the sample to




platinum-cobalt standards where one unit of color being that pro-




duced by 1 mg/1 platinum, in the form of the chloroplatinate ion.




Interference




     7.  Even a slight turbidity interfers with the test.  The




recommended method for the removal of turbidity is by centrifugation.




     The color is extremely pH dependent and invariably increases as




the pH of the sample is raised.  The pH must be measured and recorded




along with the color.




Reagents




     8.  Same as given under "Preparation of Standards", pages 128 and




129, Standard Methods for the Examination of Water and Wastewater,




Twelfth Edition, 1965.




Procedure




     9.  Same as given on page 129, Standard Methods for the Examination




of Water and Wastewater. Twelfth Edition, 1965.




Calculation




     10.  Same as given on page 129, Standard Methods for the Examination




of Water and Wastewater. Twelfth Edition, 1965.

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet  Parameter No.  00095

Parameter    :   Conductivity,  Specific (micromhos/cm at 25  C)



Method       :   Instrumental,  wheatstone bridge modified.
Reference    :   j.   Standard Methods for the Examination of Water and
                    Wastewater, 12th ed., APHA, Inc., N.Y., 1965,
                    280-284.

                2.   ASTM Book of Standard, Part 23, 1967
                    D1125-64, pp. 179-184.


Modifications:   Test conducted at standard temperature of 25 C only.
                                                Date:  September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No. to-he—assigned
Parameter    :  Copper,  Dissolved
Method
Filtration through 0.45 y membrane filter.
Measurement by atomic absorption.
Reference
                Method is specific to the particular instrument.
                General procedure is being selected.
Modifications:  Use of 0.45 y membrane filter only.
                                                Date:  September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No.  01040
Parameter    :   Copper,  Total
Method
Atomic absorption.
Reference    :   Method is  specific to the particular instrument.
                General procedure is  being selected.
Modifications:  None
                                               Date;  September  1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
Parameter
                                         Storet Parameter No.  00720
Cyanide (mg/1)
Method
Reference
Colorimetric or volumetric; distillation followed
by benzidine pyridine color development or silver
nitrate titration.

Standard Methods, 12th edition, 1965, 450-454.
Aldridge, W. N. 1944.  Analyst (69) 262-265.
Nusbaum, I and P. Skupeko, 1951, Sewage and Ind.
Wastes, 231 (7), 875.
Modifications:  Standard Methods, Page 450. Delete Section 1.1, para. 2.

                Cyanogen bromide forms an addition product with
                pyridine which in turn reacts with an aromatic
                amine with rupture of the pyridine ring to form
                an intensely colored dianilide of glutaconic
                dialdehyde.

                Benzidine is used as the aromatic amine producing
                an orange to red colored compound with maximum
                absorption at 480 my.

                See Benzidine Pyridine method write-up attached.
                                                Date;   September 1968

-------
             Federal Water Pollution Control Administration
                   Analytical Quality Control Branch
                        OFFICIAL INTERIM METHOD
                                  for
Benzidine Pyridine Method        Cyanide
Reagents
Purity of Reagents
     Reagents shall be reagent grade and should conform to all speci-
fications of the Committee on Analytical Reagents, ACS.  Other grades
of chemicals may be used if it has been determined that the reagent
is of sufficient quality to allow its use without reducing the precision
or accuracy of the method.
     Purity of Water - Unless otherwise indicated, references to water
shall be understood to mean reagent water conforming to specifications
of reagent water (ASTM D1193).
     (a)  Phosphoric Acid (85%) Concentrated phosphoric acid.
     (b)  Phosphoric Acid Solution (1:9) Mix 1 volume of H3P04 (85%)
with 9 volumes of water.
     (c)  Methyl orange indicator (0.5g/liter).  Dissolved 0.05 grams
of methyl orange in water and dilute to 100 ml.
     (d)  Bromine water (Saturated). To 250 ml of water add slightly
more liquid bromine than will dissolve on shaking.  Store in a glass
stoppered actinic glass bottle.
                                                                    SEP     1968

-------
     (e)  Sodium Arsenlte Solution - Dissolve 2 grains of Sodium




Arsenite Na2Ag03 in 100 ml of water.




     (f)  N-butyl alcohol reagent grade.




     (g)  Pyridine Solution - Combine 25 ml of pure pyridine with




2 ml of cone, hydrochloric acid, dilute to 100 ml with water.




     (h)  Benzidine Hydrochloride Solution - Dissolve 2 grams of




benzidine dihydrochloride in 100 ml of water and filter through




filter paper.




     (i)  Benzidine Pyridine mixture - Add 0.3 ml of benzidine solution




(h) to 5 ml of pyridine solution (g), mix and add immediately to sample.






Colorimetric Procedure





Take an aliquot or 20 ml of the absorption liquid obtained from the




distillation procedure.  The aliquot should contain less than 0.5




micrograms of cyanide.




     Place the aliquot in a test tube 25 x 150 or larger preferably




a screw cap culture tube; bring volume up to 20 ml with distilled.  Add




2 drops of a methyl orange solution; shake and add dropwise a 10%




phosphoric acid solution until a red color develops.  Add saturated




bromine water until definite bromine light yellow bromine color




develops.  Discharge bromine color with dropwise addition of sodium




arsenite solution and add one drop in excess.




     Add 10 ml of N-butyl alcohol.




     Mix 0.3 ml of the benzidine reagent with 5 ml of pyridine solution;




add to sample and mix by vigorous shaking.
                                                                     SEP     1968

-------
        Allow 15 minutes for color to develop,  decant carefully the

   butyl alcohol phase and measure absorbance of the developed color with

   photometer at 48Dm>i.   Determine the concentration of the cyanide in the

   sample by comparing with standards run simultaneously or when appropriate

   number of points  are  plotted  on calibration curve to permit calcula-

   tion of slope which will not  differ by one pat in a hundred.  In all

   cases at  least three  standards  are to be run each day samples are

   analyzed.


   Colorimetric  Procedure (for CN   Concentration lower than Img per liter).

        Calculate the  cyanide concentration as follows:
                                               r   Y
        CN-milligrams  per  liter  = —	——:	—r*	^r~r——	
                                 ml  original  sample x ml of aliquot used
        C  = micrograms  of  cyanide  determined.

References;

Aldridge, W. N. "A New Method for Estimation of Micro Quantities of  Cyanide
and the Cyanate,Analyst 69,262 (1944)

Nusbaum I and Skupeko Pe "Determination Cyanides in Sewage and Polluted
Waters "Sewage and Industrial Wastes 23, 7, 875 (July 1951)

American Society of Testing Materials, ASTM, Designation: DII93
                                                                  SEP    1968

-------
                               Dissolved Oxygen
                          INTERIM OFFICIAL PROCEDURE
                   Azide Modification of lodometric Method
                            Full Bottle Technique


1.  Scope and Application:
     This method is applicable for use with most wastewaters and streams that
contain nitrite nitrogen and more than 1 mg/1 of ferrous iron.  Other reducing
or oxidizing materials should be absent.  If 1 ml fluoride solution is added
before acidifying the sample and there is no delay in titration, the method is
also applicable in the presence of 100-200 mg/1 ferric iron.
     The azide modification is not applicable under the following conditions:
(a) samples containing sulfite, thiosulfate,polythionate, appreciable quantities
of free chlorine or hypochlorite (b) samples high in suspended solids (c) samples
containing organic substances which are readily oxidized in a highly alkaline
solution, or which are oxidized by free iodine in an acid solution (d) domestic
sewage (e) biological floes (f) where sample color interferes with endpoint de-
tection.
     In instances where the azide modification is not applicable a DO probe will
be used.
2.  Apparatus;
     2.1) Sample Bottles - 300 ml +_ 3 ml capacity BOD incubation bottles with
tapered ground glass pointed stoppers and flared mouths.
     2.2) Pipets - with elongated tips capable of delivering 2.0 ml _+ 0.1 ml
of reagent.
                                                            SEP    1968

-------
3.  Sampling:




     3.1) Where possible,  collect the sample in a 300 ml BOD incubation bottle.




Special cautions are required to avoid entrainment or solution of atmospheric




oxygen or dissolution of dissolved oxygen.




     3.2) Where samples are collected from shallow depths (less than 5 feet) use




of an APHA-type sampler is recommended.  Use of a Kemmerer-type sampler is




recommended for samples collected from depths of greater than 5 feet.




     3.3) When a Kemmerer is used, the BOD sample bottle should be filled to




overflowing.  (Overflow for approximately 10 seconds).  Outlet tube of Kemmerer




should be inserted to bottom of BOD bottle.  Care must be taken to prevent tur-




bulence and the formation of bubbles when filling bottle.




     3.4) The sample temperature should be recorded at time of sampling as pre-




cisely as required.




4.  Preservation of Samples:




     4.1) Do not delay the determination of dissolved oxygen in samples having




an appreciable iodine demand or containing ferrous iron.  If samples must be




preserved either 4.1.1 or 4.1.2 below may be employed.




        4.1.1) Add 2 ml of manganous sulfate reagent and then 2 ml of alkali




azide reagent to the sample contained in the BOD bottle.  Both reagents must be




added well below the surface of the liquid.  Stopper the bottle immediately and




mix the contents thoroughly.  The sample should be stored at the temperature of




the collection water; or water sealed and kept at a temperature of 10 - 20°C, in




the dark.




        4.1.2) Add 0.7 ml of concentrated ISO  and 1 ml sodium azide solution
                                                                       1953

-------
(2gNaN3 in 100 ml distilled water) to the sample in the DO bottle.   Store




in sample as in method A.   Complete the procedure using 2 ml of manganous




sulfate solution, 3 ml alkali iodide solution,  and 2 ml of concentrated




H2S04.




     4.2)  If either preservation technique A or B is employed, complete




the analysis within 4-8 hours after sampling.




5.   Reagents




     5.1)  Manganese sulfate reagent:  As in Standard Methods	




     5.2)  Alkaline-iodide-azide reagent:  As in Standard Methods	




     5.3)  Sulfuric acid,  concentrated:  As in Standard Methods	




     5.4)  Starch solution:  Prepare an emulsion of 10 g soluble starch in




a mortar or beaker with a small quantity of distilled water.  Pour this emulsion




into 1 liter of boiling water, allow to boil a few minutes, and let settle




overnight.  Use the clear supernate.  This solution may be preserved by the




addition of 5 ml per liter of chloroform and storage in a 10°C refrigerator.




Dry, powdered starch indicators such as "thyodene" may be used in place of




starch solution.




     5.5)  Potassium Fluoride solution:  Dissolve 40 g KF • 21^0 in distilled




water and dilute to 100 ml.




     5.6)  Sodium thiosulfate stock solution, 0.75N:  Dissolve 186.15 g Na^S^Oo




5H20 in boiled and cooled distilled water and dilute to 1 liter.  Preserve by




adding 5 ml chloroform.




     5.7)  Standard sodium thiosulfate titrant, 0.0375N:  Prepare by diluting




50.0 ml of stock solution to 1 liter.  Preserve by adding 5 ml of chloroform.




Standard sodium thiosulfate, exactly 0.0375N is equivalent to 0.300 mg of DO




per 100 ml.  Standardize with 0.0375N potassium dichromate.





                                                               SEP     1963

-------
     5.8)  Standard potassium dichromate, 0.0375N:  Dissolve 1.839 potassium

dichromate, previously dried 2 hours at 103°C, in 1 liter of distilled water.

The solution must be prepared in a volumetric flask.

     5.9)  Standardization of 0.0375N sodium thiosulfate:  Dissolve 2 g +

0.1 g KI in 100 to 150 ml distilled watery add 10 ml of 10% H2S04 followed

by 20 ml standard potassium dichromate.   Place in dark for 5 minutes, dilute

to 400 ml and titrate with the standard sodium thiosulfate titrant to a pale

straw color.  Add 1-2 ml starch solution and continue the titration drop by

drop until the blue color disappears.  Run in duplicate.  Duplicate deter-

mination should agree within 0.05 ml.

6 .   Procedure :

     6.1)  To the sample collected in the BOD incubation bottle add 2 ml of

the manganous sulfate solution followed by 2 ml of the alkali-iodide-azide

reagent, well below the surface of the liquid; stopper with care to exclude

air bubbles and mix well by inverting the bottle several times.  When the

precipitate settles, leaving a clear supernatant above the manganese hydroxide

floe, shake again.  With estuarine-type waters, a minimum 2 minute period of

contact with the precipitate rather than settling is sufficient.  When

settling has produced at least 100 ml of clear supernate, carefully remove

the stopper and immediately add 2.0 ml of cone. l^SO^ (sulfamic acid packets,
                                  2
3 gm, may be substituted for l^SO^ ) by allowing the acid to run down the
 The volume of standard dichromate to be employed in the standardization
 procedure may be reduced to 9 ml to accomodate the use of a 10 ml burette.

o
 Kroner, R.C., Longbottom,  J.E.,  Gorman, R. ,  A Comparison of Various Reagents
 Proposed for Use in the Winkler Procedure for Dissolved Oxygen, PHS Water
 Pollution Surveillance System Applications  and Development Report #12, Water
 Quality Section, Basic Data Branch, July, 1964.

-------
neck of the bottle,  re-stopper, and mix by gentle inversion until the iodine


is uniformly distributed throughout the bottle.  Complete the analysis within


45 minutes.


     6.2)  Transfer the entire bottle contents by inversion into a 500 ml


wide mouth erlenmeyer and titrate with 0.0375N thiosulfate solution (where

                                                                      2
problems of stability arise, 0.0375N PAO may be substituted as titrant ) to


a pale straw color.   Add 1-2 ml of starch solution or 0.1 g of powdered


indicator and continue to titrate to the first disappearance of the blue


color.


     6.3)  If ferric iron is present (100 to 200 ppm), add 1.0 ml of KF


solution before acidification.


     6.4)  Occasionally a dark brown or black precipitate persists in the


bottle after acidification.   This precipitate will dissolve if the solution


is kept for a few minutes longer than usual or, if particularly persistant,


a few more drops of l^SO^ will effect dissolution.


7.  Calculation:


     7.1)  Each ml of 0.0375 sodium thiosulfate titrant is equivalent to 1 mg/1


DO when the entire bottle contents are titrated.


     7.2)  If the results are desired in milliliters of oxygen gas per liter


at 0°C and 760 mm pressure,  multiply mg/1 DO by 0.698.


     7.3)  To express the results as percent saturation at 760 mm atmospheric


pressure, the solubility data in Table 25 (Whipple & Whipple Table, Page 409


Standard Methods, 12th Edition) may be used.  Equations for correcting the


solubilities to barometric pressures other  than mean sea  level are given


below the Table.


2
 Kroner, R.C., Longbottom, J.E., Gorman R., A Comparison  of Various Reagents

 Proposed  for Use in the Winkler Procedure  for Dissolved  Oxygen, PHS Water

 Pollution Surveillance System Applications and Development Report #12, Water

  Quality Section, Basic Data  Branch, July 1964.                  (
                                                               •

-------
     7.4)  The solubility of DO in distilled water at any barometric

pressure, P (mm Hg),  temperature,  T°C,  and saturated vapor pressure,

u (mm Hg),  for the given T,  may be calculated between the temperature of

0° and 30°C by:

                       ml/1  DO = (P - U) x 0.678
                                      35 +T

                       and between 30° and 50°C by:

                       ml/1  DO = (P - U) x 0.827
                                      49 +T

8.   Precision and Accuracy:

     Exact data are unavailable on the precision and accuracy of this

technique however, reproducibility is approximately 0.2 ppm of DO at the

7.5 ppm level due to equipment tolerances and uncompensated displacement

errors.
                                                          SCP

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No.  00950
Parameter    :   Fluoride  (mg/1)
Method
:   Colorimetric, SPADNS, with distillation.
Reference
      Standard Methods for the Examination of Water
      and Wastewater, 12th ed., APHA, Inc., N.Y., 1965,
      144-146.

      ASTM Book of Standards, Part 23, 1967
      1968 preprint, D1179.
Modifications:   None
                                               Date;  September  1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No.   00950
Parameter    :   Fluoride (mg/1)
Method
:   Colorimetric, SPADNS, with distillation.
Reference
      Standard Methods for the Examination of Water
      and Wastewater, 12th ed., APHA, Inc., N.Y., 1965,
      144-146.

      ASTM Book  of Standards, Part 23, 1967
      1968 preprint, D1179.
Modifications:  None
                                               Date;  September  1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet Parameter No.   00910

Parameter    :  Hardness,  Calcium (mg/1



Method       :  Volumetric,  with EDTA
Reference    :   ASTM Book of Standards,  Part  23,  1967
                D1126-67, pp.  188-191.
Modifications:  None
                                                Date;  September  1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No.   nnQ20
Parameter    :  Hardness, Magnesium (mg/1
Method
                Volumetric, with EDTA, by difference.
Reference
                ASTM Book of Standards, Part 23, 1967
                D1126-67, pp. 188-191.
Modifications:  None
                                                Date:   September 1968

-------
Parameter
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet Parameter No.  00900
             :  Hardness, Total (mg/1 CaC03)
Method
:  Volumetric, with EDTA.
Reference    :  ASTM Book of Standards, Part 23, 1967
               D1126-67, pp. 188-191.
Modifications:  None
                                                Date;  September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No. to be assigned
Parameter    :   Iron,  Dissolved
Method       :   Filtration  through  0.45 y membrane  filter.
                Measurement by atomic  absorption.
Reference    :  Method  is  specific  to  the particular instrument.
               General procedure is being selected.
Modifications:  Use  of  0.45 y membrane  filter  only.
                                                Date;  September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No.   0.1045
Parameter    :   Iron,  Total
Method
Atomic Absorption
Reference    :   Method is  specific  to  the  particular  instrument,
                General procedure is being selected.
Modifications:  None
                                                     September  1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No.  to be  assigned
Parameter    :   Lead,  Dissolved
Method       :   Filtration through. 0.45  y membrane  filter.
                Measurement by atomic absorption.
Reference    :   Method is  specific to the  particular  instrument.
                General procedure is  being selected.
Modifications:  Use of 0.45 y membrane filter  only.
                                                Date:  September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No.  01049
Parameter    :   Lead,  Total
Method
                Atomic absorption.
Reference    :   Method is  specific  to  the particular  instrument.
                General procedure is being selected.
Modifications:  None
                                                Date;  September  1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No.  to be assigned
Parameter    :   Magnesium,  Dissolved
Method       :   Filtration through 0.45 y  membrane filter.
                Measurement by atomic absorption.
Reference    :   Method is specific to the particular instrument.
                General procedure is being selected.
Modifications:  Use of 0.45 y  membrane filter only.
                                                Date t

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control' Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No.  00925
Parameter    :  Magnesium, Total
Method
             :  Atomic absorption.
Reference    :  Method is specific to the particular instrument.
                General procedure is being selected.
Modifications:  None
                                               Date; September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No.  to be assigned
Parameter    :   Manganese,  Dissolved
Method       :   Filtration through 0.45 y membrane filter.
                Measurement by atomic absorption.
Reference    :   Method is specific to the particular instrument.
                General procedure is being selected.
Modifications:  Use of 0.45 y membrane filter only.
                                                Date:  September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No.  01055
Parameter    :   Manganese, Total
Method
             :   Atomic  absorption.
Reference    :   Method is  specific  to  the particular instrument.
                General procedure is being  selected.
Modifications:   None
                                                Date:  SePtember 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet Parameter No.  38260

Parameter    :  MBAS (Methylene Blue Active Substance)  (mg/1)
Method
Methylene Blue Method.
Reference    :  Standard Methods for the Examination of Water and
                Wastewater,  12th ed., APHA,  Inc.,  N.Y., 1965,
                297-299.
Modifications:  None
                                                Date;   September 19.68

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No.   00610
Parameter    :  Nitrogen, Ammonia  (mg/1)
Method       :   Manual  - Distillation.
                Automated - In preparation.
Reference    :  i.   Standard Methods for the Examination of Water and
                    Wastewater, 12th ed., APHA, Inc., N.Y., 1965,
                    187-193.

                2.   ASTM Book of Standards, Part 23, 1967
                    D1426-58, pp. 336-338.


Modifications:  New  method write-ups. See attached.
                                                Date:  September 1968

-------
             Federal Water Pollution Control Administration




                   Analytical Quality Control Branch




                        OFFICIAL INTERIM METHOD




                                  for




                    Nitrogen-Ammonia by Distillation







1.  Scope




     1.1  This distillation method covers the determination of ammonia-




nitrogen exclusive of total Kjeldahl nitrogen in surface waters (both




fresh and saline) and industrial waste water.




2.  Purity of Reagents




     2.1  Reagent grade chemicals shall be used in all tests.  Unless




otherwise indicated, it is intended that all reagents shall conform




to the specifications of the Committee on Analytical Reagents of the




American Chemical Society, where such specifications are available.




Other grades of reagents may be used, provided it is first ascertained




that the reagent is of sufficiently high purity to permit its use




without reducing the accuracy of the determination.




     2.2  All reagents and standards are to be prepared with ammonia—free




water. Such water is best prepared by the passage of distilled water through




an ion-exchange column containing a strongly acidic cation-exchange resin




mixed with a strongly basic anion-exchange resin.  These resins should




also be selected so that organic compounds which might subsequently




interfere with the ammonia determination will be removed.
                                                                          196?

-------
                                                                  2.




3.  Sample Preservation




     3.1 Until more conclusive data are obtained samples may be




preserved by the addition of 1.0 ml concentrated I^SO^ or 40 mg Hg"1"1"




per liter.  If possible, the samples should be kept in a frozen state




until analyzed.






                         DISTILLATION METHOD




4.  Application




     4.1  This method is applicable to all types of water and is




the method of choice where economics and sample load do not warrant




the use of automated equipment.




5.  Summary of Method




     5.1  The sample is buffered at a pH of 9.5 wi±h a borate buffer




in order to decrease hydrolysis of cyanates and organic nitrogen




compounds, and is then distilled into a solution of boric acid. The




ammonia in the distillate can be determined either colorimetrically




by nesslerization or titrimetrically with standard sulfuric acid with




the use of a mixed indicator, the choice between these two procedures




depending on the concentration of the ammonia.  In the analysis of sea




water the calcium and magnesium salts remain in solution as their




soluble borates.




6. " Interferences




     6.1  A number of aromatic and aliphatic amines, as well as other




compounds, both organic and inorganic, will cause turbidity upon the




addition of Nessler reagent, so direct nesslerization has been discarded
                                                                  SEP
                                                                         1968

-------
                                                                  3.





as an official method.  Cyanate, which may be encountered in certain




industrial effluents, will hydrolyze to some extent even at the pH




of 9.5 at which distillation is carried out.  Volatile alkaline




compounds such as hydrazine will influence the titrimetric results.




Some volatile compounds, such as certain ketones, aldehydes, and




alcohols, may cause an off-color upon nesslerization in the dis-




tillation method.  Some of these, such as formaldehyde, may be




eliminated by boiling off at a low pH prior to distillation and




nesslerization.  Residual chlorine must also be removed by pre-




treatment of the sample with sodium thiosulfate before distillation.





7.  Apparatus




     7.1  An all-glass distilling apparatus with an 800-1000 ml




flask.




     7.2  Spectrophotometer or filter photometer for use at 425 my




and providing a light path of 1 cm or more.




     7.3  Nessler tubes: Matched Nessler tubes (APHA Standard)




about 300 mm long, 17 mm inside diameter, and marked at 225 mm +_




1.5 mm inside measurement from bottom.





     7.4  Erlenmeyer flasks:  The distillate is collected in 500 ml




glass-stoppered flasks.  These flasks should be marked at the 350 and




the 500 ml volumes.  With such marking, it is not necessary to




transfer the distillate to volumetric flasks.
                                                                    SEP     1968

-------
                                                                  4.

8.  Reagents

     8.1  Ammonium chloride, standard solution. - (1 ml =

0.01 mg NH3-N). - Dissolve 3.819 g NH^Cl in water and bring

to volume in a 1 liter volumetric flask for use as a stock solution.

Dilute 10 ml of this stock solution to 1 liter in a volumetric flask

for use as the standard ammonium chloride solution.

     8.2  Boric acid solution (20 g/1). - Dissolve 20 g H3B03 in

water and dilute to 1 liter.

     8.3  Mixed indicator. - Mix 2 volumes of 0.2 percent methyl

red in 95 percent ethyl alcohol with 1 volume of 0.2 percent methylene

blue in 95 percent ethyl alcohol.  This solution should be prepared

fresh every 30 days.

     Note 1. — Specially denatured ethyl alcohol conforming to
     Formula 3A or 30 of the U.S. Bureau of Internal Revenue may
     be substituted for 95 percent ethanol.

     8.4  Nessler reagent. - Dissolve 100 g of mercuric iodide and

70 g of potassium iodide in a small amount of water.  Add this mixture

slowly, with stirring, to a cooled solution of 160 g of NaOH in 500 ml

of water.  Dilute the mixture to 1 liter.  If this reagent is stored

in a Pyrex bottle out of direct sunlight, it will remain stable for a

period of up to 1 year.

     Note 2. — This reagent should give the characteristic color
     with ammonia within 10 minutes after addition, and should not
     produce a precipitate with small amounts of ammonia (0.04 mg in
     a 50 ml volume).

     8.5  Borate buffer. - Add 88 ml.  of 0.1 N NaOH solution to 500 ml

of 0.025 M sodium tetraborate solution (.5.0 g Na^B^O^ per liter) and

dilute to 1 liter.
                                                                      SEP

-------
                                                                  5.

     8.6  Sulfuric acid,  standard solution (0.02 N, 1 ml =

0.28 mg NH3~N).  - Prepare a stock solution of approximately 0.1 N acid

by diluting 3 ml of concentrated 1^504 ^SP 8r I-8*) to 1 liter with

C02~free distilled water.  Dilute 200 ml of this solution to 1 liter

with C02-free distilled water.  Standardize the approximately 0.02 N

acid so prepared againstO.0200 N Na2C03 solution.  This last solution

is prepared by dissolving 1.060 g anhydrous Na2CC>3, oven-dried at 140° C,

and diluting to 1 liter with (X^-free distilled water.

     Note 3. — An alternate and perhaps preferable method is to
     standardize the approximately 0.1 N l^SO^ solution against a
     0.100 N Na2C03 solution.  By proper dilution  the 0.0200 N acid
     can then be prepared.

     8.7  Sodium hydroxide (IN). - Dissolve 40 g NaOH in ammonia-free

water and dilute to 1 liter.

     8.8 Dechlorinating reagents. - A number of dechlorinating reagents

may be used to remove residual chlorine prior to distillation.  These

include:

     (a) Sodium thiosulfate  (1/70 N):  Dissolve 3.5 g Na2S203 in

ammonia-free water and dilute to 1 liter.  One ml  of this solution

will remove 1 mg/1 of residual chlorine in 500 ml  of sample.

     (b)  Sodium arsenite (1/70 N):  Dissolve 1.0  g NaAs02 in ammonia-

free water and dilute to 1 liter.

9.  Procedure

     9.1  Preparation of equipment.  — Add 500 ml  of ammonia-free

water to an 800 ml Kjeldahl  flask.   The addition of boiling  chips

which have been previously treated with dilute NaOH will prevent

bumping.  Steam out the distillation apparatus until the distillate

shows no trace of ammonia with Nessler reagent.


                                                               SEP    1968

-------
                                                                  5.




     9.2  Sample preparation. - To 400 ml of sample add 1.0 N NaOH




until the pH is 9.5, checking the pH during addition with a pH meter.




     9.3  Distillation. - Transfer the sample, the pH of which has been




adjusted to 9.5, to an 800 ml Kjeldahl flask and add 25 ml of the




borate buffer.  Distill 300 ml at the rate of 6-10 ml/min. into 50 ml




of 2% boric acid contained in a 500 ml glass stoppered Erlenmeyer flask.




Dilute the distillate to 500 ml in the flask and nesslerize an aliquot




to obtain an approximate value of the ammonia—nitrogen concentration.  For




concentrations above 1.0 mg/1 the ammonia should be determined titri-




metrically.  For concentrations below this value it is determined




colorimetrically.




     9.4  Determination of ammonia in distillate. - Determine the ammonia




content of the distillate either titrimetrically or colorimetrically




as described below.  (See 9.4.1 and 9.4.2).




     9.4.1  Titrimetric determination. — Add 3 drops of the mixed




indicator to the distillate and titrate the ammonia with the 0.02 N




H2s<-*4» notching the end point against a blank containing the same




volume of ammonia-free water and 113603 solution.




     9.4.2  Colorimetric determination. — Prepare a series of 14




Nessler tubes containing the following volumes of the standard solution:




Q.0,,0.2, 0.4, 0.6, 0.8, 1.0, 1.2, 1.4, 1.7, 2.0, 2.5, 3.0, 3.5, and 4.0




ml.  Dilute each tube to 50 ml. with ammonia-free water.  Add 1 ml of




Nessler reagent and mix.  After 20 minutes read the optical densities at




425 my against a compensating blank of nesslerized ammonia—free water.




From the values obtained prepare a standard curve.
                                                                      SEP

-------
                                                                  6.




     Determine the ammonia in the distillate by nesslerizing 50 ml




or an aliquot diluted to 50 ml and reading the optical density at




425 my as described above for the Standards. Ammonia-nitrogen content




is read from the standard curve.




10.Calculations




     (Tentative until a Standard Method for calculating results is




adopted).




     10.1 Titrimetric



             /i x™  XT     A x 0.28 x 1000
           mg/1 NH3-N  =	





           in which:




           A  =  ml 0.02 N H2S04 used




           S  =  ml sample




     10.2  Spectrophotometric



             II XTU  XT     A X 1000
           mg/1 NH3-N  =   Q>8 s





           in which:




           A  =  mg NHo^ read from standard curve




           S  =  volume of distillate nesslerized




11.  Precision




     11.1  Precision and accuracy data are not available at this time.
                                                                        SEP     1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet Parameter No.  QQ625


Parameter    :  Nitrogen, Kjeldahl, Total  (mg/1)
Method
Reference
               Acid digestion and distillation, with titration
               or nesslerization, modified.
               Standard Methods for the Examination of Water and
               Wastewater, 12th ed., APHA, Inc., N.Y., 1965,
               403-404.
Modifications: See method write-up.
                                                Date:   September 1968

-------
      Federal Water Pollution Control Administration

            Analytical Quality Control Branch

                  OFFICIAL INTERIM METHOD

                  Hitrogen-Kjeldahl Total

Scope

1.  This method covers the determination of total Kjeldahl
nitrogen in fresh and estuarine waters.  Two alternatives
are listed for the determination of ammonia after distillation
in the official method.  The Kjeldahl procedure converts the
nitrogen of most components of biological origin, such as
amino acids, peptides and proteins, but may not convert the
nitrogeneous matter in many industrial wastes, such as
azines, hydrazones, nitro compounds, oximes, semi carbazones,
and some refractory tertiary amines.  The titrimetric method
is applicable to concentrations above 2 mg N per liter and
possibly to lower concentrations if reagent blanks are con-
sistent enough.  The nesslerization method is applicable to
concentrations above 0.1 mg N per liter.  The lower limit
can be extended by concentration of the distillate prior to
nesslerization.  While the official method is outlined for a
macro system, the same chemistry applied in a micro system
(Standard Methods for the Examination of Water and Waste-
water, llth Edition, pg 305) is considered to be the same as
the official method.

Definitions

?.  (a) Total Kjeldahl nitrogen is defined as the sum of free
ammonia and of organic nitrogen compounds which are converted
to (NH^^SO^ under the conditions of digestion which are
specified below.

    (b) Organic Kjeldahl nitrogen is defined as the differ-
ence obtained by subtracting the free ammonia value (reference
to be cited) from the total Kjeldahl nitrogen value.  This may
be determined directly by removal of ammonia before digestion.

    (c) Organic nitrogen is defined as the nitrogen derived
from all organic substances in the sample and shall include
Kjeldahl nitrogen.
                                                   SEP    1963

-------
                                                        2.1
Purity of Reagents
3.  (a) Reagent grade chemicals shall be used in all tests.
Unless otherwise indicated, it is intended that all reagents
shall conform to the specifications of the committee on
Analytical Reagents of the American Chemical Society, where
such specifications are available.  Other reagents may be
used, provided it is first ascertained that the reagent is
of sufficiently high purity to permit its use without
lessening the accuracy of the determination.  Sulfuric acid,
in particular, is subject to absorption of ammonia and should
be protected from possible ammonia contamination.

    (b) vJater shall be understood to mean distilled water
which is free of ammonia.  Such water is best prepared by
the passage of distilled water through an ion-exchange column
containing a strongly acidic cation-exchange resin mixed with
a strongly basic anion-exchange resin.  These resins should
be selected so that organic compounds which might subsequently
interfere with the determination will be removed.  (Note 1)

    Note 1:  The regeneration of the ion-exchange column
             should be carried out according to the
             instruction of the manufacturer.

4.  Sampling and Preservation

    (a)  Collect the samples in accordance with the appli-
         cable methods recommended by the Committee on
         Sampling.

    (b)  The samples can be preserved for analysis at ice
         temperature for a period not exceeding 48 hours.
         Samples can be preserved with 1 ml cone. H^SO^
         per liter of sample and retained for longer periods
         provided that the relationship between organic
         nitrogen and ammonia is not of interest.  The
         H^SO^ will retain ammonia in solution; however,
         it accelerates the conversion of organic nitrogen
         to ammonia.  Samples treated with H^SO^ are also
         subject to gain ammonia from the air if not
         tightly sealed.
                                                     SEP    1963

-------
                                                      2.2


                   Digestion Method
Application
5.   This method is applicable to all types of fresh and
estuarine waters, containing from 0 to 100 mg N per liter.

Summary of Method

6.  The sample is heated in the presence of concentrated
sulfuric acid, K^SO^ and HgO and evaporated until SO^ fumes
are obtained and the solution becomes colorless or pale
yellow.  The residue is cooled, diluted and is treated and
made alkaline with a hydroxide-thiosulfate solution.  The
ammonia is distilled and determined after distillation either
by nesslerization (see appropriate references) or titrimet-
rically.

Apparatus

7.  (a)  Digestion Apparatus

    Kjeldahl digestion apparatus with 800 ml flasks and suction
takeoff to remove water and S0~ fumes.

    (b)  Distillation Apparatus

    The Kjeldahl flask is connected to a condenser and the
adaptor is so arranged that the distillate can be collected
for nesslerization or in indicating H-jBO-j solution for titration.

    (c)  Colorimeter of Nessler tubes as required for nessler-
ization (see ammonia nitrogen).

Reagents

8.  (a)  Mercuric Sulfate Solution

    Dissolve 8 g red mercuric oxide, HgO, in 50 ml 1:5
H90-H7S04 solution and dilute to 100 ml with H?0.
                                                    SEP    1S68

-------
      (b)  Sulfuric Acid - Mercuric Sulfate - Potassium
Sulfate Solution

     Dissolve 267g ^SO* in 1,300 ml water and add 400 ml
concentrated VSO.  Add 50 ml mercuric sulfate solution and
dilute to 9. liters.

     (c)  Sodium Hydroxide - Sodium Thiosulfate Solution

     Dissolve 500g HaOH and 25g Na2S203 5H20 in water and
dilute to 1 liter.

     (d)  Phenolphthalein Indicator Solution

     Dissolve 5g phenolphthalein in 500 ml 95% ethyl alcohol
or isopropanol and add 500 ml water.  Add 0.02 NaOH dropwise
until a faint pink color appears.

     (e)  Mixed Indicator

     Mix 2 volumes of 0.270 methyl red in 95% ethanol with 1
volume of 0.270 methylene blue in ethanol.  Prepare fresh every
30 days.

     (f)  Boric Acid Solution

     Dissolve 20g boric acid, H^BO^, in water and dilute to
1 liter with water.

     (g)  Sulfuric Acid Titrant, 0.020N.

     In this strength 1.00 ml = 0.28 mg N.

Procedure

9.   (a)  Preparation of Apparatus

     The distillation apparatus should be presteamed before
use by distilling a 1:1 mixture of ammonia-free water and
sodium hydroxide -sodium thiosulfate solution until the
distillate is ammonia free.  This operation should be
repeated each time the apparatus is out of service long
enough to accumulate ammonia (usually 4 hours or more).

     (b)  Digestion

     Place a measured sample or the residue from the distil-
lation in the ammonia determination (for Organic Kjeldahl only)
into an 800 ml Kjeldahl flask.   The sample size can be deter-
mined from the follox^ing table:

                                                    SEP    1958

-------
0 -
5 -
10 -
20 -
50 -
5
10
20
50
100
          Kjeldahl Nitrogen          Sample Size
           in sample, mg/1           	ml	

                                        500
                                        250
                                        100
                                         50.0
                                         25.0

Dilute the sample, if required, to 500 ml, and add 100 ml
sulfuric acid - mercuric sulfate - potassium sulfate solution
(Note 2), and evaporate the mixture in the Kjeldahl apparatus
until 363 fumes are given off and the solution turns colorless
or pale yellow.  Cool the residue and add 300 ml water.

     Note 2:  Alternately digest the sample with 1 Kel-Pac
              (Olin-Matheson) and 20 ml l^SO^.

     (c)  Distillation

     Make the digestate solution alkaline by carefully adding
sodium hydroxide - thiosulfate solution without mixing.  Connect
a condenser to the Kjeldahl flask and connect the apparatus
in such a manner that the tip of the condenser extends well
below the level of 100 ml of boric acid.  Mix and distill
until 200 ml of distillate is collected.  Add 1 ml of mixed
indicator and titrate with 0.0200N H^SC^ to a pale lavender
endpoint.  Run a blank on the reagents and correct as required.

Calculation

10.  (a)  Calculate total Kjeldahl nitrogen, in milligrams per
liter, in the original sample as follows:

     total Kjeldahl nitrogen, mg/1 - (A-B) x N x F x 1000
                                               S
     or A-B x 280
     where:

     A = milliliters of standard 0.020N H?SO, solution used
     in titrating sample

     B = milliliters of standard 0.020N H?SO, solution used
     in titrating blank

     N = normality of sulfuric acid solution
                                                   SEP
1963

-------
     F = millequivalent weight of nitrogen (14 mg)

     3 = milliliters of sample digested

     (b)  If nesslerization is used, follow ammonia
procedure calculations.

     (c)  Calculated organic Kjeldahl nitrogen, in
milligrams per liter, as follows:

     organic Kjeldahl nitrogen, mg/1 - C - D

     whe re:

     C = total Kjeldahl nitrogen in mg/1

     D = ammonia nitrogen, in mg/1

Precision and Accuracy

11.  The precision and accuracy will be established by
FWPCA laboratories.
                                                 SEP    1963

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet Parameter No.  00620

             :   Nitrogen, Nitrate (ing/1)


                Manual Method: Brucine Sulfate, colorimetric for fresh
             :                  and salt water.
                Automated Method A: reduction, colorimetric for fresh water.
                Automated Method B: Brucine Sulfate, colorimetric for
                               saline water.  Cto be added at later date)

                1.  Standard Methods for the Examination of Water and
                    Wastewater, 12th ed., APHA, Inc., N.Y., 1965,
                    198-200.

                2.  ASTM Book of Standards, Part 23, 1967,
                    D992-52, pp. 138-140.

Modifications:   New method write-ups.  See attached.
Parameter
Method
Reference
                                                Date:  September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                        OFFICIAL INTERIM METHOD

                                  for

                            Nitrogen-Nitrate

1.  Scope

     1.1  The following methods are suitable for the determination

of nitrogen-nitrate in water samples:

     a.  Manual - Brucine Sulfate (fresh and salt water)

     b.  Automated A, Hydrazine Reduction (fresh water)

     c.  Automated B, Brucine Sulfate (saline water).
           (.to be added at later date)

2.  Purity of Reagents

     2.1  Reagent grade chemicals shall be used in all tests.  Unless

otherwise indicated, it is intended that all reagents shall conform

to the specifications of the Committee on Analytical Reagents of the

American Chemical Society where such specifications are available.

Other grades of reagents may be used, provided it is first ascertained

that the reagent is of sufficiently high purity to permit its use

without reducing the accuracy of the determination.  Distilled water

free of nitrite and nitrate is to be used in preparation of all reagents

and standards.

3.  Sample Preservation

     3.1  Until more conclusive data is obtained, samples may be

preserved as follows:

     (a)   Addition of 40 mg Hg/1 (as HgC^)  and stored at 4° C.

     (b)   Addition of 1 ml of CHC13/100 ml sample.   Store at 4°  C.
                                                                  SEP    1968

-------
                                                                  2.

     (c)   Freezing.

           Note 1.  — Data available at the present time indicate
           that in using the above methods the nitrate concentration
           is stable over a period of at least six weeks.
                            MANUAL METHOD

4.  Application and Principle of Method

     4.1  This method is applicable to a wider variety of samples than

either of the two automated methods, because modification can be made

to remove or correct for turbidity, color, salinity, or dissolved

organic compounds in the samples.

     4.2  The principle of the method is based upon the reaction of the

nitrate ion with brucine sulfate in a 13 N t^SO^ solution at about 100° C.

The optical density of the resulting colored complex is measured at

410 my.

     4.3  Applicable range is 0.1 to 2 mg N-NO-/1.

5.  Interferences

     5.1  Dissolved organic matter will cause an off color in 13 N

H^SO^ and must be compensated for by additions of all reagents except

the brucine-sulf anilic acid reagent.  This also applies to natural

color present not due to dissolved organics.

     5.2  If turbidity is present, it must be removed by prior filtration,

preferably through a 0.45 y membrane filter.

     5.3  The effect of salinity is eliminated by addition of sodium

chloride to the blanks, standards and samples.

     5.4  All strong oxidizing or reducing agents interfere.  The presence

of oxidizing agents may be determined by the addition of orthotolidine

reagent.

-------
                                                                  3.




     5.5  Residual chlorine interference is eliminated by the




addition of sodium arsenite.




     5.6  Ferrous and Ferric iron and quadrivalent manganese give




slight positive interference, but in concentrations less than 1 mg/1




these are negligible.




6.  Apparatus




     6.1  Spectrophotometer or filter photometer suitable for measuring




optical densities at 410 my and capable of accommodating 25 mm diameter




cells.




     6.2  Sufficient number of 25 mm diameter matched tubes for




reagent blanks, standards, and samples.




     6.3  Neoprene coated wire racks to hold 25 mm diameter tubes.




     6.4  Water bath suitable for use at 100° C.  This bath should




contain a stirring mechanism so that all tubes are at same temperature.




     6.5  Water bath suitable for use at 10-15°C.




7.  Reagents




     7.1  Sodium Chloride Solution  (300 g/1).  - Dissolve 300 g NaCl




in distilled water and dilute to 1000 ml.




     7.2  Sulfuric Acid Solution. - Carefully add 500 ml H2S04 (SP 8r 1-84)




to 125 ml distilled water.  Cool and keep tightly stoppered to prevent




absorption of atmospheric moisture.




     7.3  Brucine-Sulfanilic Acid Reagent. - Dissolve 1 g brucine




sulfate I(C23 H26 N^^-l^SCy 7 H20] and 0.1 g sulfanilic acid




(NH2 C6 H4 S03H . H20) in 70 ml hot distilled water.  Add 3 ml concentrated




HC1, cool, mix and dilute to 100 ml.  Store in a dark bottle at 5° C.
                                                                  SEP     1968

-------
                                                                  4.




This solution is stable for several months; the pink color that develops




slowly does not effect its usefulness.  Mark bottle with warning;




   CAUTION; Brucine jSulfate is toxic; take care to avoid ingestion.




     7.4  Stock Potassium Nitrate Solution  (1 ml  =  0.1 mg N03~N)




Dissolve 0.7218 g anhydrous KNO, in distilled water and dilute to 1 liter,




     7.5  Standard Potassium Nitrate Solution  (1ml  =  0.01 mg




N03-N). - Dilute 100 ml of the stock solution to 1 liter. This standard




solution should be prepared fresh weekly.




     7.6  Acetic Acid (1+3) - Dilute 1 voL glacial CH  COOH with




3 vol. distilled water.




8.  Procedure




     8.1  Adjust the pH of the samples to approximately pH 7 with




1+3 acetic acid and if necessary filter through a 0.45 y membrane




filter or Whatman No. 42 filter paper depending on the amount and




type of suspended solids present.




     8.2  Set up the required number of matched tubes in the rack to




handle reagent blank, standards and samples.  If it is necessary to




correct for color or dissolved organic matter (which will cause color




on heating), a set of duplicate tubes must be used to which all reagents




except the brucine-sulfanilic acid has been added.




     8.3  Pipet 10 ml or an aliquot of the samples diluted to 10 ml




into the sample tubes.




     8.4  If the samples are saline, add 2.0 ml of the 30 percent




NaCl solution to the reagent blank, standards and samples.  For fresh




water samples NaCl may be omitted.  Mix contents of tubes of swirling




and place rack in cold water bath (0-10°C).
                                                                   SEP     1968

-------
                                                                   5.



      8.5   Pipet  10 ml  of H2^A solution  into  each  tube and mix by



 swirling.  Allow tubes to  come to  thermal equilibrium in  the cold  bath.



      8.6   Add 0.5 ml brucine-sulfanilic  acid  reagent to each tube



 (except the  interference control tubes)  and carefully mix by swirling,



 then  place the rack of tubes in the boiling water  bath for exactly



 25 minutes.



      8.7   Remove rack  of tubes from the  hot water  bath and immerse



 in the cold  water bath and allow to reach thermal  equilibrium (20-25°C).



      8.8   Dry tubes and read optical density  against the  reagent blank



 at 410 my.



 9.  Calculation



      9.1   Obtain a standard curve by plotting the  optical densities



 of standards run by the above procedure  against mg NO--N.  The color



 reaction does not always follow Beer's law.



      9.2   Subtract the  absorbance of the sample without the brucine-



 sulfanilic reagent from the absorbance of the sample containing



 brucine-sulfanilic acid and read the resultant absorbance in mg NO--N.



 Multiply by  factor for converting mg per aliquot of sample to mg per liter.





     mg/1 NO.,-N   =     100° .    X mg NO,-N from curve
            3       ml sample     &   3





                         AUTOMATED METHOD
10.  Application



     10.1  This method is applicable to water samples which contain less



than 500 mg/1 calcium.  The applicable range of this method is 0.05-10 mg/1



nitrite or nitrate nitrogen.  Approximately 20 samples per hour can be



analyzed.




                                                                    SEP

-------
                                                                  6.

11.  Summary of Method

     11.1  This procedure, using the Technicon Auto Analyzer, determines

NO--N by the conventional diazotization—coupling reaction.  The NO--N

is reduced with hydrazine sulfate in another portion of the sample and

the nitrite thus formed is determined in the usual manner.

     11.2   Subtraction of the NO^-N originally present in the sample

from the total NO»-N found will give the original NO--N concentration

in terms of N02-N.

12.  Interferences

     12.1  The following table lists the concentration of ions that

cause no interference in the determination of nitrite and nitrate

nitrogen.  The same interfering ion concentration applies to either

nitrite or nitrate:

              Ion                 Mg/1 ion not causing interference

              Cl~                            30,000

              P04~3                              50


              S~2                              Note

              NH3-N                              80

              Mg+2                               75

              Ca+2                              240

              Fe+3                               30

              ABS                                60

     Note 2. —  The apparent NO^ and NOo concentrations varied +
     10 percent with concentrations of sulfide ion up to 10 mg/1.

     12.2  The pH of the samples should be between 6 and 9.
                                                                      SEP     1968

-------
                                                                  7.

13.  Apparatus

     13.1    Technicon Auto Analyzer consisting of:

     13.1.1  Two proportional pumps.

     13.1.2  Two colorimeters each with an 8 mm flow-through cell and

520 my filters.

     13.1.3  One continuous filter.

     13.1.4  One sampler (large).

     13.1.5  Two recorders.

     13.1.6  One 38° C temperature bath.

     13.1.7  Two time delay coils.

14.  Reagents

     14.1  Color developing reagent.  - To approximately 3 liters of distilled

water add 400 ml concentrated phosphoric acid (sp gr 1.834),  60 g

sulfanilamide (H^N-CgH  S02 NH2) followed by 3.0 g N(l-Naphthyl)ethylene-

diamine dihydrocholoride.  Dilute the solution to 4 liters with distilled

water and store in a dark bottle in the refrigerator.  This solution is

stable for approximately 1 month.

     Note 3. — It may be necessary to apply heat in order to
     dissolve the sulfanilamide.

     14.2  Copper sulfate; stock solution. - Dissolve 2.5 g of copper

sulfate (CuSO^ • 5 H20) in distilled water and dilute to 1 liter.

     14.3  Copper sulfate; dilute solution. - Dilute 20 ml of stock

solution to 2 liters with distilled water.

     14.4  Sodium hydroxide; stock solution. (10 N). - Dissolve 400 g

NaOH in 750 ml distilled water, cool and dilute to 1 liter.
                                                                     SEP     1968

-------
                                                                  8.




     14.5  Sodium hydroxide (.1.0 N) .  - Dilute 100 ml of stock NaOH




solution to 1 liter.




     14.6  Sodium hydroxide CO.3 N) .  - Dilute 60 ml of stock NaOH




to 2 liters.




     14.7  Hydrazine sulfate solution. - Dissolve 54.92 g of hydrazine




sulfate (N2H  •  H SO.) in 1800 ml of distilled water and dilute to




2000 ml.  This solution is stable for approximately 6 months.




          CAUTION:  Toxic if ingested.  Mark container with warning.




     14.8  Nitrate Standards




     14.8.1  Potassium nitrate; stock standard (1000 mg/1 • NO--N).




Dissolve 7.2180 g of KNO-, oven dried at 100-105° C for 2 hours, in




distilled water and dilute to 1000 ml.  Add 1 ml chloroform as a




preservative.  Stable for six months.




     14.8.2  Potassium nitrate, dilute standard (100 mg/1 NO--N).




Dilute 50 ml of stock KNO, solution to 500 ml in a volumetric flask.




From this dilute solution prepare the following standards in 500 ml




volumetric flasks:




               mg/1 N03~N            ml standard solution




                   0.4                       2.0




                   1.0                       5.0




                   1.6                       8.0




                   3.0                      15.0




                   5.0                      25.0




                   7.0                      35.0




                  10.0                      50.0
                                                                    SEP     1968

-------
                                                                  9.




     14.9  Nitrite Standards




     14.9.1  Potassium nitrite, stock standard O-OOO mg/1 NO--N).




Dissolve 4.9260 g NaN02, oven dried at 100-105° C for two hours, in




distilled water and dilute to 1000 ml.  Add 1 ml chloroform as




preservative. Store in the refrigerator.  Stable for one month.




     14.9.2  Potassium nitrite, dilute standard (100 mg/1).




Dilute 50 ml of stock NaNO~ solution to 500 ml in a volumetric flask.




From this dilute solution prepare the same volumetric standards as




in 15.3.  Prepare fresh each week.




15.  Procedure




     15.1  Set up the manifold as shown in Figures 1 and 2. Allow




both colorimeter (with the proper filters) and recorder to warm up




for 30 minutes, then run a base line with all reagents, feeding distilled




water through the sample line.  Adjust dark current and operative




opening on each colorimeter.  Adjust base line to 0.01 optical density.




Place a distilled water wash tube in alternate openings on sampler




and set sample timing at 1.5 minutes.




     15.2  Run a 2.0 mg/1 NO--N and a 2.0 mg/1 N02~N standard through




the system to check for 100% reduction of nitrate to nitrite.  The two




peaks should be of equal height.  If the NO- peak is lower than that




of the N0£ peak, the temperature of the reduction bath should be




increased until they are equal.  If the NO- peak is higher than the




nitrate, the temperature should be reduced.  When the correct temperature




of the bath has been determined, no further adjustment should be necessary.
                                                                  SEP    196?

-------
                                                                  10.




     15.3  Arrange standards in sampler in NO--NO- order with increasing




concentration of nitrogen.  Place unknown samples in sampler tubes and




place in alternate openings of sampler.  A NO- and NO- standard of equal




nitrogen concentration should be placed after every 10 samples as a




further check on the system and to more easily identify peaks.




16.  Calculation




     16.1  Subtract the NO- concentration in the sample from the total




NO- found (nitrite plus nitrate) on the reduction side to calculate




the NO- concentration in the sample.




17.  Precision




     17.1  Precision and accuracy data are not available at this time.






                              References




1.  D. Jenkins and L. Medsker, "Brucine Method for Determination of




    Nitrate in Ocean, Estuarine, and Fresh Waters".  Anal. Chem., 36.




    610 (1964).




2.  L. Kamphake, S. Hannah, and J. Cohen, "Automated Analysis for




    Nitrate by Hydrazine Reduction".  Water Research, l> 205 (1967).
                                                                 SEP

-------
                                          Figure 1.   Nitrate-Nitrite Manifold
                                                                                               August 1968
                        Large Mixing Coil
Double
Delay
Coil
Continuous Filter
                       Large Mixing Coil
                        From Figure 2
                        for N0« Determination

                                                               Black   Black
   ml/min.
   2.90 Sample
                                                                               1-20 Air
   0.32 N NaOH
   0.42 N NaOH
                               SAMPLER I
£


£•
                                                                               2.50 Waste from NO^ Colorimeter
                                                                               2.50 Waste from N02 Colorimeter
                                                               PROPORTIONING
                                                                  PUMP
                                     HS
                Sampling Time:   1.5 Min
                Wash Tubes   :   One
                                            COLORIMETER     RECORDER
                                          8 mm Tubular  f/c
                                        520 mu Filters

-------
                                             Figure 2.  Nitrate-Nitrite Manifold
                                                                                                     August  1968
o
u

60
C
• r-l
X
•r-l
S

01
00
"for N02 Determinatiot
^
f 	
\i/ \L
Large Mixing Coil
, OOOOQMQ X
f M JS f
'PL L..,^
/^\\ n°^!e
1 fc^/ Coil


Z Waste.
R
Y
P
0
Black
G
G
Blue
Y
R
Y
B
0
Black
G
G
Blue
Y

Blue
Blue
Blue
Blue
PROPORTIONING
PUMP

f

gggSSJ j '71 J^
T
SSL 	 1 ^ iT


ml/min.
„ 0.80 Waste
_ 1.20 Air
^2.90 Dist. H20
" 0.42 N02Sample ' ^
IFrom Filter
(See Figure 1)
\ f
^2.00 Cu Reagent
^2.00 0.30 N NaOH
H 1.60 Hydrazine Reagent
^ 1.20 Air

^ 1.60 Dist. H20
^ 1.60 Color Reagent
To N02 Manifold
Proceed to Figure 1
Sampling Time: 1.5 Min.
Wash Tubes : One
                                               COLORIMETER
                                            8 mm Tubular  f/c

                                          520 mn Filters
RECORDER

-------
Method
Reference
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No. 00615
Parameter    :   Nitrogen, Nitrite  (mg/1)
Manual Method: colorimetric diazotization, modified.
Manual Automated Method: same as manual.
Automated Method: included in automated nitrate method
    for fresh water.

1.  Standard Methods for the Examination of Water and
    Wastewater, 12th ed., APHA, Inc.,  N.Y., 1965,
    205-208.

2.  ASTM Book of Standards, Part 23,  1967
    D1254-67, pp. 256-260.
Modifications:  New method write-up.  See  attached.
                                                Date:  September  1968

-------
      Federal Water Pollution Control Administration

            Analytical Quality Control Branch

                  OFFICIAL INTERIM METHOD

                           for

                     Nitrogen-Nitrite


Scope
1.        These methods are applicable for the determina-

          tion of nitrite ion in water, wastewater, sewage,

          sludges, and saline water by colorimetric methods.

          Official Method - Colorimetric method diazota-

          tion with sulfanilamide and N-(l-naphthyl)-

          ethylenediamine.

          Official Method - Automated - same chemistry as

          manual - to be developed for .010 tng/1 sensitivity.*

     Preservation

?..        Samples should be preserved at time of collection

          with 1 ml of chloroform per 100 ml sample or 40 mg

          Hg/ml as mercuric chloride.  Store at 4°C.  DO NOT

          PRESSRV2 WITH SULFURIC ACID.

     Summary of Method

3.        The diazonium compound formed by diazotation of

          sulfanilamide by nitrite ion in water under acid

          conditions is coupled with N-(l-naphthyl)-


     *Automated nitrite method is included in the automated

      nitrate method for fresh waters.


                                                   SEP    1368

-------
          ethylenedtamine to produce a reddish-purple color



          which is read in a spectrophotometer at 520 mn.



     Interferences



4.        There are very few known interferences at concen-



          trations less than 1000 times that of the nitrite;



          however, recent addition of strong oxidant or



          reductants to the sample will readily affect the



          nitrite concentration.



Determination of Nitrite-Nitrogen Ion in Aqueous Medium



     Reagents



5.   a)   Color reagent:  To 400 ml of distilled water, while



          stirring on a magnetic mixer, add 50 ml cone. HCl,



          5.0 g of sulfanilamide and 0.5 g of N-(l-naphthyl)-



          ethylenediamine dihydrochloride.  Stir until



          dissolved and dilute to 500 ml.



     b)   Sodium acetate buffer, 2M:  To 50 ml, while stirring



          on a magnetic mixer, add 27.4 g of NaC2Ho09.3 H^O.



          Stir until dissolved and dilute to 100 ml.



     c)   Stock nitrite-nitrogen standard (100 mg/1



          Dissolve 0.4926 g sodium nitrite, anhydrous



          and dilute to 1000 ml with distilled water.



          Preserved with 2 ml chloroform per liter.  Stable



          one week.
                                                   SEP    1968

-------
     Apparatus

6.   a)   Spectrophotometer or filter photometer suited for

          use in range of 515 to 530 tnp.

     b)   Nessler tubes of 50 ml volume.

     Calibration

7.   a)   Prepare in 50 ml Nessler tubes  a reagent blank and

          five standards in the desired concentration range

          from the 100 mg/1 NC^-N standard and dilute to

          about 40 ml.  The slope of the  curve should be

          checked each day with a blank and one standard.

     b)   To each add 1.0 ml of color reagent and 1.0 ml of

          acetate buffer.  Dilute to 50 ml and allow color

          to develop for 10 minutes.  The color reaction

          medium should be between pH 2.0 and ?.5.

     c)   An estimate of the color density which will permit

          approximately 25% transmission in various size

          cells can be obtained from the  following:


           ml stock   mg/50 ml   mg/1   cell length
           standard    N09-N     N02-N      cm
2.0
4.0
20.0
.002
.004
.020
.040
.080
.40
10.
5.
1.
     d)   clead in Spectrophotometer at 520 mn against

          reagent blank.
                                                SEP    !S68

-------
      Procedure

 8.   a)   If the sample has a pH greater than 10, adjust

           it to approximately pH 7 with acetic acid (1:3).

      b)   Filter sample through Whatman No.  42 paper or

           .45 (a membrane filter, depending on the amount

           of suspended solids in sample.

      c)   To 40 ml (or aliquot diluted to 40 ml) add 1.0 ml

           color roagent and 1.0 ml of buffer.  Dilute to

           50 ml mark and mix.

      d)   Read color after 10-30 minutes in spectrophotometer

           at 520 mp against reagent blank in same size cells

           as used with nitrite-nitrogen standards.

      Calculations

 9.   a)   The colorimetric reaction obeys Beer-Lambert's

           law so that in plotting the absorbancy of the

           standards against the concentration, a straight

           line results.

      b)   Calculate the concentration of nitrite-nitrogen

           ion in the sample in milligram per liter as follows:

           mg/1 NC>2-N = absorbance of sample x mg/1 standard x	
                                               absorbance standard
           1000
           ml sample

      Precision

10.        To be evaluated by "round robin" testing.
                                               SEP    1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No.  OQ3QQ
Parameter    :  Oxygen Dissolved  (mg/1)
Method
:  Volumetric, Winkler azide, modified.
Reference    :   1.  Standard Methods for the Examination of Water and
                   Wastewater, 12th ed., APHA, Inc., N.Y., 1965.
                   406-410.

                2.  ASTM Book of Standards, Part 23, 1967
                   D888, pp. 114-115.


Modifications:   Full bottle technique with normality adjusted
                to 0.0375N sodium thiosulfate. See new method
                write—up.
                                                Date:  September 1968

-------
Parameter
Federal Water Pollution Control Administration

      Analytical Quality Control Branch

          OFFICIAL INTERIM METHODS

                     for

      Analyses of Surface Water Samples


                            Storet Parameter No.  00666

:   Phosphorus, Soluble (mg/1)
Method
   Colorimetric, mixed reagent, filtered with persulfate
   digestion.
Reference
Modifications:
   1.  Gales, M. E. Jr., Julian, E. C., and Kroner, R. C.,
       "Method for Quantitative Determination of Total
       Phosphorus in Water." J. AWWA, .58(10), 1363-1368
       (1966).

   2.  Murphy, J. and Riley, J. P., "A Modified Single
       Solution Method for the Determination of Phosphate
       in Natural Waters," Anal. Chim. Acta, 27. 31-36 (1962),

   New method write-up.
                                               Date;  September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet Parameter No.  to be assigned

Parameter    *  Phosphorus, Soluble, Ortho (mg/1)
Method
Colorimetric, mixed reagent,  filtered.
Reference
Murphy, J. and Riley, J. P., "A Modified Single
Solution Method for the Determination of Phosphate
in Natural Waters." Anal. Chim. Acta, 27. 31-36 (1962)
Modifications:  New method write-up.
                                                Date:   September  1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
Parameter
                                         Storet Parameter No. 00665
             :   Phosphorus, Total (mg/1)
Method
:   Colorimetric,  mixed reagent,  with persulfate  digestion
Reference
Modifications:
   1.  Gales,  M.  E.  Jr.,  Julian,  E.  C.,  and Kroner, R.  C.,
       "Method for Quantitative Determination of Total
       Phosphorus in Water." J. AWWA,  j>8(10), 1363-1368
       (1966).

   2.  Murphy, J. and Riley, J. P.,  "A Modified Single
       Solution Method for the Determination of Phosphate
       in Natural Waters." Anal.  Chim. Acta, .27, 31-36 (1962)

   New method  write-up.
                                                Date;  September 1968

-------
Parameter
Federal Water Pollution Control Administration

      Analytical Quality Control Branch

          OFFICIAL INTERIM METHODS

                     for

      Analyses of Surface Water Samples


                            Storet Parameter No.  to be assigned

:   Phosphorus, Total, Ortho (mg/1)
Method
   Colorimetric, mixed reagent.
Reference
   Murphy, J. and Rlley, J. P., "A Modified Single
   Solution Method for the Determination of Phosphate
   in Natural Waters." Anal. Chim. Acta, 2]_t 31-36 Q-962).
Modifications:   New method write-up.
                                                Date:  September 1968

-------
                                                                August 1968
                   Analytical  Quality Control  Branch
                     FWPCA OFFICIAL INTERIM METHOD
                              PHOSPHORUS
I.  Scope
    I.I  These methods cover the determination of specified forms of
         phosphorus in surface waters, saline  waters,  industrial  waters,
         and sewage-type water samples.   They  may be applicable to
         sediment-type samples,  sludges, algal  blooms  etc., but
         sufficient data is not available at this time to warrant such
         usage when measurements for total  phosphorus  content are
         requi red.
    1.2  The methods are based on reactions that are specific for the
         orthophosphate ion.  Thus, depending  on the prescribed pre-
         treatment of the sample, the various  forms of phosphorus
         given in Figure I  may be determined.   These forms are, in
         turn, defined in Table I.
         1.2.1  Except for in-depth and  detailed studies, the most
                commonly measured forms  of phosphorus  are total and
                soluble phosphorus, and  total  and soluble ortho-
                phosphate.   Hydro IyzabIe phosphorus is normally found
                only in sewage-type samples and insoluble forms of
                phosphorus, as noted, are determined by calculation.
    1.3  The methods are usable in the 0.01  to 0.5 mg/I  P range.
2.  Definitions
    2.1  The various forms  of  phosphorus are defined in Table I.
                                                          SEP     1358

-------
Residue
                       SAMPLE

                             Filter
(No Fi It rat- inn")
                                               Direct
                                               Colorimetry
                                                Total
                                            Orthophosphate
                   H2S04
                   Hydrolysis &
                   Colorimetry
                 Total Hydrolyzable
                  & Orthophosphate
\
Direct
Colorimetry
Soluble
Orthophosphate

J
Hydrolysis
& Colorimetry
Soluble Hydrolyzable
& Orthophosphate

\
Persulfate
Digestion
& Colorimetry
Total Soluble
Phosphorus
                                                                                             \
  Persulfate
  Digestion
  & Colorimetry
  Total
Phosphorus
        Figure 1.  Analytical Scheme For Differentiation of Phosphorus Forms.
                                                                                                               ^953

-------
                             -3-










                           TABLE  I





                   PHOSPHORUS TERM INOLOGY








I.   Total  Phosphorus (P-T)  -  all  of  the  phosphorus  present in  the



    sample,  regardless of  form, as  measured  by  the  persulfate



    digestion procedure.




    a.  Total  Orthophosphate  (P-T,ortho) -  inorganic phosphorus




        dCPO.)  3 in the  sample as  measured  by  the  direct color-




        imetric analysis  procedure.



    b.  Total  Hydrolyzable Phosphorus (P-T,hydro) - phosphorus in




        the sample as measured by the suIfuric  acid hydrolysis




        procedure, and minus  pre-determined  total orthophosphates.




        This hydrolyzable  phosphorus includes polyphosphates C(PoO-,)   ,



        (P,0._)  , etc.]  + some organic  phosphorus.




    c.  Total  Organic Phosphorus  (P-T,org)  - phosphorus (inorganic +



        oxidizable organic) in the  sample  as measured by the




        persulfate digestion  procedure,  and  minus total hydrolyzable



        phosphorus and orthophosphate.




2.   Total  Soluble Phosphorus  (P-S,T) - all  of the phosphorus present




    in the filtrate of a  sample filtered through a  phosphorus-free




    filter of 0.45 micron  pore size and  measured by the persulfate




    digestion procedure.




    a.  Soluble Orthophosphate (P-S,ortho)  - as measured by the




        direct colorimetric analysis procedure.




                                                        SEP     1368

-------
                              -4-
                            lAbLE  i  (Contd.)









    b.   Soluble Hydrolyzable Phosphorus  (P-S,hydro)  - as  measured



        by the sulfuric acid hydrolysis  procedure and minus pre-




        determined  soluble  orthophosphates.




    c.   Soluble Organic Phosphorus  (P-S,org)  - as measured  by the




        persulfate  digestion procedure,  and minus soluble hydro-




        lyzable phosphorus  and orthophosphate.




3.   The following forms, when sufficient amounts of  phosphorus are




    present in the  sample to warrant such consideration,  may be




    caIculated:




    a.   Total  Insoluble Phosphorus  (P-I,T) =  (P-T) - (P-S,T).



        (I)  Insoluble Orthophosphate (P-l,ortho) =  (P-T,ortho) -




        (P-S,ortho).




        (2)  Insoluble Hydrolyzable Phosphorus (P-l,hydro)  =




        (P-T,hydro) - (P-S,hydro).




        (3)  Insoluble Organic Phosphorus (P-l,org)  = (P-T,org) -



        (P-S,org).





4.   All phosphorus  forms shall be reported as P, mg/I.
                                                          SEP     1968

-------
                             -5-
3.   Samp i i ng




    3.1   Samples should be taken in a location that is the most




         representative of the waters in the area being sampled.




         If benthic deposits are present in the area being sampled,




         great care should be taken not to include these deposits.




    3.2  Sample containers may be of plastic material, such as




         cubitainers,  or of Pyrex glass.  In both cases, high iron




         concentrations interfere through the formation of clumps




         in the bottom of the sample.



    3.3  If the analysis cannot be performed the same day of




         collection, the sample should be preserved in one of the




         following manners (Note I):




         3.3.1  Five ml of chloroform per liter of sample.




         3.3.2  Acidification with sulfuric acid (used only when




                not measuring for all forms of phosphorus).




         3.3.3  Quick-freezing of sample.




         3.3.4  Five ml of saturated solution of mercury chloride




                per gallon of sample.




                Note I  - Sufficient data is not available on these



                suggested preservatives to state a preference at




                this time.





4.   Summary of Method [£)oI orimetric Single Reagent Method]




    4.1  Ammonium molybdate and potassium antimony I tartrate react




         in an acid medium with dilute solutions of phosphorus to




                                                        SEP    1968

-------
                           -6-
         form an antimony-phosphate-moiybdate complex.   This




         complex is reduced to an intensely blue-colored complex



         by ascorbic acid.  The color is proportional  to the



         phosphorus concentration.




    4.2  Only orthophosphate forms  a blue color in this test.




         Polyphosphates (and some organic phosphorus compounds)




         may be converted to the orthophosphate form by suIf uric-




         acid-hydrolysis.  Organic  phosphorus compounds may be



         converted to the orthophosphate form by persulfate




         digestion.





5.  Interferences



    5.1  It is reported   that no interference is caused by




         copper, iron, or silicate  at concentrations many times



         greater than their greatest reported concentration in




         sea water.



    5.2  The salt error for samples ranging from 5 to 20 percent




         salt content was found to  be less than I  percent




    5.3  Arsenate, in concentrations greater than found in sea water,




         does not interfere   :
                                                                    1963

-------
                                    —7-
                                 TABLE  2
             Effect of  Other Ions  on  Determination  of  Phosphate
Concentration Optical density at 822 my
Ion
—
Copper( 1 1 )
1 ron ( 1 1 1 )
Sil icate
Arsenate
Arsenate
mg/l
^
10
50
10
1.0
0.02
No added P04/P
0.012
0.018
0.017
0.014
1 . 1 44a
0.020
+ 10 ygP04/P
0.593
0.585
0.598
0.594
I.7523
0.614
(4-cm eel Is)
Di f ference
0.581
0.567
0.581
0.580
0.608
0.594
a
 Measured in I-cm cell  calculated for 4-cm cells.


     6.  Apparatus

         6.1  Photometer - A spectrophotometer or filter photometer suitable

              for measurements at 880 my,  and providing a light path of I

              inch (2.54 cm) or longer,  should be used.

         6.2  Acid-washed glassware:   All  glassware used in the determination

              should be washed with hot 1:1 HCI  and rinsed with distilled

              water.  The acid-washed glassware should be filled with

              distilled water and treated  with all the reagents to remove

              the last traces of phosphorus that might be adsorbed on the

              glassware.  Preferably, this glassware should be used only for

              the determination of phosphorus and after use it should be

              rinsed with distilled water and kept covered until needed again..

              If this  is done, the treatment with  1:1 HCI and reagents is

              only required occasionally.   Commercial detergents should never be

              used.
                                                                     StP     1363

-------
                         -8-
7.1   Sulfuric acid  solution,  5N:   Dilute  70  ml  of  cone.  H-SO.



     with distilled water  to  500  ml.



7.2  Potassium antimony I tartrate solution:   Weigh 1.3715 g



     KCSbOC.H.O... 1/2  H00,  dissolve  in  400 ml  distilled  water
            446     Z.


     in 500 ml  volumetric  flask,  dilute to volume.  Store in



     glass-stoppered bottle.



7.3  Ammonium molybdate solution: Dissolve  20 g  (NH.),.Mo7074.4 hLO



     in 500 ml  distilled water.   Store  in a  plastic bottle at  4° C.



7.4  Ascorbic acid, O.IM:   Dissolve  1.76  g of ascorbic acid  in



     100 ml  of distilled water.   The solution is  stable  for



     about a week if stored at 4° C.



7.5  Combined reagent:  Mix the above reagents in  the following



     proportions for 100 ml of the mixed  reagent:   50 ml of



     5N HUSO., 5 ml of potassium  antimony I tartrate solution,  15



     ml of ammonium molybdate solution, and  30 ml  of ascorbic



     acid solution. Mix after addition of each reagent.  All



     reagents must  reach room temperature before  they are mixed



     and must be mixed  in  the order  given.   If turbidity forms



     in the combined reagent, shake  and let  it stand for a few



     minutes until  the turbidity  disappears  before proceeding.



     The reagent is stable for one week if stored  at 4°  C.



7.6  Primary phosphorus standard: Dissolve  in distilled water



     0.2197 g of potassium dihydrogen phosphate,  KH_PO.   which
                                                   £.  4,




                                                      SEP    1958

-------
                              —9—
         has been dried  in  an  oven at 105°  C.   Dilute.the  solution

         to 1,000 ml;  1.00  ml  equals  0.05 mg  P.

    7tl  Working phosphorus standard:  Dilute 10.0  ml  of stock

         phosphorus solution to 1,000 ml with distilled water;

         I.00 ml equals  0.5 yg P.

    7.8  Strong-acid solution:  Slowly add  310 ml cone. H-SO.  to

         600 ml  distilled water.   When cool,  dilute to I  liter.

    7.9  Ammonium persulfate.


8.  CaIibration

    8.1   Standard curve:  Prepare  a series  of at least nine standard

         phosphorus solutions  to cover the  range from 0 to 0.50  P

         mg/l.  Make up  the standards by diluting suitable volumes

         of working phosphorus solution to  50 ml  with distilled  water.

         The following dilutions are  suggested:

                  ml of  Working
               Phosphorus Standard         Cone.,  mg/l

                       0                       0.00
                     1.0                       0.01
                     3.0                       0.03
                     5.0                       0.05
                    10.0                       0.10
                    20.0                       0.20
                    30.0                       0.30
                    40.0                       0.40
                    50.0                       0.50

    8.2  Develop color in the  series  of standards and blank of dis-

         tilled  water as directed  in  9. Procedure.  Measure the

         color absorbance of each  standard  at 880 my with  a spectro-

         photometer, using  the reagent blank  as the reference solution.

                                                              SEP     1963

-------
                            -10-
         P!ot the absorbance values as  ordinates  and  the  corre-




         sponding phosphorus concentrations  as  abscissas.



    8.3  Process standards and blank exactly as the samples.




         Run at least a blank and two standards with  each  series




         of samples.   If the standards  do not agree within ±  2%




         of the true  value,  prepare a new calibration curve.





9.  Procedure



    9.1  Total  Phosphorus




         9.1.1   Add  I  ml  of  strong-acid solution  to a 50-ml




                sample in a  125-ml  Erlenmeyer flask.



         9.1.2  Add 0.4 gram of ammonium persulfate.




         9.1.3  Boil  gently  on a pre-heated  hot plate for




                approximately 30-40 minutes  or  until  a final




                volume of about 10 ml is reached. Do not allow




                sample to go to dryness.  Alternatively,  heat




                for 30 minutes in an autoclave  at 121° C  (15-




                20 psi).



         9.1.4  Cool  and dilute the sample to 50  ml.




         9.1.5  The sample is now ready for determination of




                phosphorus as outlined  in 9.3 Orthophosphate.



    9.2  Hydrolyzable Phosphorus




         9.2.1   Add  I  ml  of  strong-acid solution  to a 50-ml




                sample in a  125-ml  Erlenmeyer Flask.




                                                      SEP    1S58

-------
         9.2.2  Boil gently on a pre-heated hot plate for




                approximately 30-40 minutes or until a final



                volume of about  10 ml is reached.  Do not allow




                sample to go to dryness.  Alternatively, heat




                for 30 minutes in an autoclave at  121° C (15-




                20 psi).



         9.2.3  Cool and dilute the sample to 50 ml.




         9.2.4  The sample  is now ready for determination of




                phosphorus  as outlined  in 9.3 Orthophosphate.





     9.3  Orthophosphate




         9.3.1  Add I drop  of phenolphthalein  indicator to  the




                50.0 ml sample.   If a red color develops, add




                strong-acid solution drop-wise to  just discharge



                the color.




         9.3.2  Add 8.0 ml  of combined  reagent to  sample and mix




                thoroughly.  After a minimum of ten minutes, but




                no  longer than thirty minutes, measure the  color




                absorbance  of each sample at 880 my with a  spectre-




                photometer, using the reagent  blank as the  reference




                solution.





10.   Calculation



     10.I   Obtain  concentration value of sample directly  from prepared




           standard  curve.   Report results as P, mg/I.





II.   Precision and Accuracy




     I I.I   Complete  precision and  accuracy data are not available
                                                             SEP
                                                                    1SG3

-------
                        -12-
      at this  time.   In a comparative study of three modifi-




      cations  of  the  Single Reagent Method,   however, the




      following results were obtained.




      I I.I.I   On  a  variety of  natural water samples, both




              salt  and fresh,  a  precision of  I.10%  (coefficient




              of  variation) and  an accuracy of 0.43$ (deviation




              from  amount added)  were obtained at a 0.2 mg/l  P




              concentration.







                      References
I.   J.  Murphy  and  J.  Riley,  "A  Modified  Single  Solution Method




    for the Determination  of Phosphate  in  Natural  Waters."




    Anal.  Chim.  Acta.,  27_,  31  (1962).




2.   M.  Gales,  Jr., E.  Julian, and  R.  Kroner,  "Method  for




    Quantitative Determination  of  Total  Phosphorus in Water."




    Jour AWWA, 58^, No.  10,  1363 (1966).




3.   J.  Winter  and  R.  Booth,  "A  Comparison  of  Three Modifications




    of  the Single  Reagent  Method  for  Phosphorus, Ortho, Soluble."




    Analytical Quality Control  Branch,  Division of Research,




    Federal Water  Pollution  Control Administration (August  1968).
                                                        SEP

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet Parameter No.   00085

Parameter    :   Odor,  Threshold,  Hot (60 C)



Method       :   Dilution Series,  modified.
Reference    :   Standard Methods for the Examination of Water and
                Wastewater,  12th ed.,  APHA,  Inc.,  N.Y., 1965,
                304-311.
Modifications:  Method in preparation.   Use reference method until
                ready.
                                                Date:  SePtember 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet Parameter No.  to be assigned

Parameter    :  Odor,  Threshold, Room Temperature



Method       :  Dilution  Series, modified.
Reference    :   Standard Methods  for the  Examination  of Water  and
                Wastewater,  12th  ed.,  APHA,  Inc.,  N.Y., 1965,
                304-311.
Modifications:  Method in preparation.   Use reference method until
                ready.
                                               Date;  September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet  Parameter No.   00550

Parameter    :   Oil and Grease (mg/1)



Method       :      Soxhlet Extraction Method .
Reference    :    Standard Methods  for the Examination of Water and
                 Wastewater,  12th  ed. ,  APHA,  Inc.,  N.Y., 1965,
                 384-385.
Modifications:   Page 384  Delete only the heading,  "1.  General Discussion".

      Insert the following heading and paragraph:

      1.  Scope and Application

           In the determination of grease, an absolute quantity of a
      specific substance is not measured.   Rather, groups of substances
      with similar physical characteristics are determined quantitatively,
      based on their mutual solubility in the solvent used.  Grease,
      therefore, can be said to include fats, waxes, oils, and any other
      nonvolatile material extracted by hexane from an acidified sample
      of water or waste water.
           Sampling and Storage;  When possible, collect representative
      samples in a wide-mouth bottle calibrated to hold a measured volume,
      and perform the initial steps of the procedure in the sample bottle.
      When information is required concerning the average grease concentration
      of a waste over an extended period,  the examination of individual
      portions collected at prescribed time intervals can be used to
      eliminate losses of grease on sampling equipment during the collection
      of a composite sample.
                                                Date; September 1968

-------
                                                                    2.

                Federal Water Pollution Control Administration

                          OFFICIAL INTERIM METHODS



Modifications Cont'd:

                 Pa%e  385  Delete paragraph 4.1.

                 Insert as paragraph 4.1:

                 Collect sample in a wide-mouth bottle and acidify
                 pH 1.0 with cone. HC1.

                 Page  385  Section 6.   Precision and Accuracy

                 Delete paragraph entirely.

                 Insert the following sentence:

                 Precision and accuracy will be determined by round
                 robin testing.
                                                                      SEP
                                                                              1958

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet Parameter No.  00681

Parameter    :  Organic  Carbon,  Dissolved (mg/1)
Method
Dow Beckman-type organic carbon analyzer.
Reference    :  ASTM Book of  Standards,  Part  23,  1967
                D2579-67T, pp.  716-719.
Modifications:  None
                                               Date;  September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet Parameter No.  00680

Parameter    :   Organic Carbon, Total (mg/1)
Method
             :   Dow Beckman-type organic  carbon  analyzer.
Reference
             :   ASTM Book of Standards, Part  23,  1967
                D2579-67T,  pp.  716-719.
Modifications:   None
                                                Date;  September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
Parameter
                                         Storet Parameter No. QQ400
                 pH
Method
Instrumental, glass/standard calomel electrodes.
Reference
1.  Standard Methods for the Examination of Water and
    Wastewater, 12th ed., APHA, Inc., N.Y., 1965,
    226-228.
2.  ASTM Book of Standards, Part 23, 1967
    289-299.
Modifications:   None
                                                Date;  September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
Parameter
                                         Storet Parameter No.   32730
Phenolics (mg/1)
Method
Reference
"Phenolic Compounds in Industrial Water and
Industrial Waste Water
Method A. Chloroform Extraction Method
Method B. Direct Colorimetric Method"

ASTM Book of Standards, Part 23, 1967
D1783-62, pp. 430-436.
Modifications:  None
                                                Date; September  1968

-------
Method
Reference
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No. 00935
Parameter    :   Potassium,  (mg/1)
                1.   Atomic absorption.
                2.   Flame photometry.
                1.   Method is specific to the particular instrument.
                    General procedure is being selected.

                2.   Standard Methods for the Examination of Water and
                    Wastewater,  12th ed. , APHA, Inc.,  N.Y., 1965,
                    238-240.
Modifications:  None
                                                Date:  September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No. to be assigned
Parameter    :   Selenium,  Dissolved (yg/1)
Method       :  Filtration through 0.45 y membrane filter.
                Measurement by diaminobenzidine colorimetric method.
Reference    :   Standard Methods for the Examination of Water and
                Wastewater,  12th ed., APHA,  Inc.,  N.Y., 1965,
                251-253.
Modifications:  Use of 0.45 y  membrane filter only.
                                                Date;  September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No.   01145
Parameter    :   Selenium,  Total  (yg/1)
Method
             :   Diaminobenzidine,  colorimetric.
Reference    :   Standard Methods  for  the  Examination  of Water  and
                Wastewater,  12th  ed.,  APHA,  Inc.,  N.Y., 1965,
                251-253.
Modifications:  None
                                                Date;  September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet Parameter No.  to be  assigned

Parameter    :   Silicon (Silica),  Dissolved  (yg/1)
Method       :   Filtration  through  0.45 u membrane  filter.
                Measurement by molybdate blue.
Reference    :   "Methods  for  Collection  of Analysis  of Water  Samples",
                Geological  Survey Water-Supply Paper 1454,  1960,
                pp.  259-261.
Modifications:  Use of 0.45  u  membrane  filter  only.
                                                Date;   September  1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet Parameter No. 01140

Parameter    :   Silicon (Silica),  Total (yg/1)
Method
Molybdate blue
Reference    :   "Methods  for Collection of  Analysis  of Water  Samples",
                Geological Survey Water-Supply Paper 1454,  1960,
                pp.  259-261.
Modifications:  None
                                                Date:  SePtember  1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
Parameter
                                         Storet Parameter No. 00930
                 Sodium (mg/1)
Method
Reference
Modifications:
                 1.   Atomic  absorption.
                 2.   Flame photometry.
                 1.   Method  is  specific  to  the particular instrument.
                     General procedure is being  selected.

                 2.   Standard Method  for the Examination of Water  and
                     Wastewater,  12th ed.,  APHA, Inc., N.Y.,  1965,
                     274-277.
                None
                                               Date: September  1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet  Parameter No. Q0520

Parameter    :   Solids, Dissolved, Volatile  (mg/1)
Method
:  Gravimetric by difference, modified.
Reference    :   Standard Methods  for  the Examination of Water and
                Wastewater,  12th  ed., APHA, Inc., N.Y., 1965,
                247-248.
Modifications:  Filter by 0.45 y membrane  filter  for dissolved/suspended
                volatile solids.
                                                Date:  September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No. 00500
Parameter    :   Solids,  Total (mg/1)
Method       :   Gravimetric,  modified.
Reference    :   Standard Methods for the Examination of Water and
                Wastewater,  12th ed.,  APHA,  Inc.,  N.Y., 1965,
                244-245.
Modifications:  Evaporate at 103-105°C only.
                                                Date:   SePtember 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet Parameter No.   QQ515

Parameter    :   Solids,  Total Dissolved (mg/1)
Method
             :   Gravimetric,  modified.
Reference    :   Standard Methods for the Examination of Water and
                Wastewater,  12th ed.,  APHA,  Inc.,  N.Y., 1965,
                247-248.
Modifications:  Use of 0.45 y membrane filter only,  for filtrations.
                Glass pre-filter may be used to speed filtration.
                Evaporate at 103-105°C only.
                                                Date:  September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet Parameter No.   00530

Parameter    :   Solids,  Total Suspended (mg/1)
Method
             :  Gravimetric,  modified.
Reference    :   Standard Methods for the Examination of Water and
                Wastewater,  12th ed.,  APHA,  Inc.,  N.Y., 1965,
                246-247.
Modifications:   Filter by  0.45  y  membrane  filter  only,
                evaporate  at  103-105°C only.
                                               Date;   September  1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet Parameter No.  00535

Parameter    :  Solids,  Suspended,  Volatile  (mg/1)
Method
             :  Gravimetric by difference,  modified.
Reference    :   Standard Methods for the Examination of Water and
                Uastewater,  12th ed.,  APHA,  Inc.,  N.Y. , 1965,
                247-248.
Modifications:  Filter by 0.45 y membrane filter for dissolved/suspended
                volatile solids.
                                                      September 1968
                                                Date:    v

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples


                                         Storet Parameter No.  00505

Parameter    :  Solids, Total, Volatile (mg/1)
Method
•   Gravimetric, modified.
Reference    :   Standard Methods for the Examination of Water and
                Wastewater,  12th ed.,  APHA,  Inc.,  N.Y., 1965,
                247-248.
Modifications:  Filter by 0.45 y membrane filter only for
                filterable/non-filterable volatile solids.
                                                Date-  SePtember 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No.  00945
Parameter    :  Sulfate (mg/1)
Method
Turbidimetric, modified.
Reference    :   Standard Methods for the Examination of Water and
                Wastewater,  12th ed.,  APHA,  Inc.,  N.Y., 1965,
                291-293.
Modifications:  Use of proprietary reagents is acceptable.
                                                Date;  September 1968

-------
             Federal Water Pollution Control Administration




                   Analytical Quality Control Branch




                       OFFICIAL INTERIM METHODS




                                  for




                   Analyses of Surface Water Samples






                                         Storet Parameter No.   00746




Parameter    :  Sulfide,  Dissolved (mg/1)
Method
Reference
             :  In preparation.
Modifications:
                                               Date: September 1968

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No.   00010
Parameter    :  Temperature (°Centigrade)
Method       :   Thermometer,  (mercury,  dial,  or thermistor).
Reference    :   Standard Methods  for the Examination of  Water and
                Wastewater,  12th  ed., APHA,  Inc.,  N.Y.,  1965,
                433.
Modifications:  None
                                                Date:  SePtember 1968

-------
Method
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No.  00076
Parameter    :   Turbidity,  Nephelometric
:   Selected instrument such as Hach 2100, based on
   method specifications.
Reference    :   New method write-up.   See attached.
Modifications:  None
                                               Date:  September  1968

-------
             Federal Water Pollution Control Administration




                   Analytical Quality Control Branch




                      OFFICIAL INTERIM METHOD




                                 for




                              Turbidity




Method Selection




     To be measured by selected turbidimeter only.  COMVAQC, FWPCA




has developed the following specifications for a turbidimeter acceptable




for use in the Administration.  Instruments are now coming on to the




market which can meet these specifications.  No instrument listing is made




since the instrument companies make quick model and design changes and




any listing would be unintentionally incomplete and possibly incorrect.




Background




     Clear water is important in those industries where the product




is destined for human consumption or for a large number of industrial




uses.  Beverage producers, food processors, and treatment plants




drawing upon a surface water supply commonly rely on coagulation,




settling, and filtration measures to insure an acceptable effluent.




     Turbidity in water is caused by the presence of suspended matter,




such as clay, silt, finely divided organic and inorganic matter,




plankton, and other microscopic organisms.  Turbidity should be clearly




understood to be an expression of the optical property of a sample




which causes light to be scattered and absorbed rather than transmitted




in straight lines through the sample.  Attempts to correlate turbidity
                                                                           -868

-------
                                 -2-





with the weight concentration of suspended matter are impractical, as




the size, shape, and refractive index of the particulate materials are




of great importance optically but bear little direct relationship to




the concentration and specific gravity of the suspended matter.




     The standard method for the determination of turbidity has




been based on the Jackson Candle turbidimeter.  Howerver, the lowest




turbidity value which can be measured directly on this instrument is




25 units.  With turbidities of treated water generally falling within




the range 0-5 units, indirect, secondary methods have been required to




obtain turbidities on such samples.  Unfortunately no instrument has




been devised which will duplicate the results obtained on the Jackson




Candle turbidimeter for all samples.  Owing to fundamental differences




in optical systems, the results obtained with different types of




secondary instruments will frequently not check closely with one




another even though the instruments are all pre-calibrated against




the Jackson Candle turbidimeter.




     A further cause of discrepancies in turbidity analysis is the




use of suspensions of different types of particulate matter for prep-




aration of instrumental calibration curves.   As with the water




samples, prepared suspensions have  different optical properties




depending upon the particle size distributions, shapes, and refractive




indices.  Most commercial turbidimeters available for measuring low

-------
                                 -3-





turbidities give comparatively good indications  of the intensity  of




light scattered in one particular direction,  predominately at  right




angles to the incident light.   Since there is no direct relationship




between the Jackson Candle turbidity and the  intensity of light




scattered at 90°, there is no valid basis for the practice of  calibrating




the 90° turbidimeters in terms of Jackson units.




1.  General Discussion




    1.1.  Principle:  The method presented below is based upon a  comparison




of the intensity of light scattered by the sample under defined conditions




with the intensity of light scattered by a standard reference  suspension




under the same conditions.  The higher the intensity of scattered light,




the higher the turbidity.  Formazin polymer,  which has gained  acceptance




as the turbidity standard reference suspension in the brewing  industry,




is also used as the turbidity standard reference suspension for water.




It is easy to prepare and is more reproducible in its light scattering




properties than the clay or turbid natural water standards previously




used.  The turbidity of a particular concentration of Formazin suspension




is defined as 40 units.  This same suspension of Formazin has  an




approximate turbidity of 40 units when measured on the Jackson Candle




turbidimeter, therefore turbidity units based on the Formazin
                                                                 SEP     1963

-------
                                -4-
prcparation will approximate those derived  from the  Jackson  Candle




turbidimeter but will not be identical to  them.




    1.2.   Interference:   The determination  of  turbidity  is applicable




to any water sample that is free of debris  and coarse  sediments  which




settle out rapidly, although dirty glassware,  the  presence of air




bubbles,  and the effects of vibrations which disturb the surface




visibility of the sample will lead to false results.   The presence  of




"true color", that is the color of the water which is  due to dissolved




substances which absorb light, will cause measured turbidities to be




low.  This effect is generally not significant with treated  water.




    1.3.   Storage:  It is preferable to determine  turbidity  on the  same




day the sample is taken.  If longer storage is unavoidable,  however,




samples may be stored in the dark up to 24  hr.  For even longer storage,




treat 1 liter of sample with Ig mercuric chloride.  Prolonged storage




before measurement is not recommended as irreversible changes in




turbidity may occur.  All samples should be vigorously shaken before




examination.






2.  Apparatus




    2.1  The turbidimeter shall consist of a nephelometer with light




souroe for illuminating the sample and one or  more photoelectric




detectors with a readout device to indicate the intensity of  light




scattered at right angles to the path of the incident light.  The




turbidimeter should be so designed that little stray light  reaches  the




detector in the absence of turbidity and should  be free from significant




drift after a short warm-up period.  The sensitivity of the  instrument




should permit detection of turbidity differences of 0.02 unit or less




in waters having turbidities less than 1 unit.  The instrument should  SEP    1963

-------
                               -5-
mcasurc. from 0-40 units turbidity.   Several ranges will be necessary




to obtain both adequate coverage and sufficient sensitivity for low




turbidities.




       The sample tubes to be used  with the available instrument must be




of clear, colorless glass.  They should be kept scupulously clean,




both inside and out, and discarded  when they become scratched or etched.




They must not be handled at ail where the light strikes them, but




should be provided with sufficient  extra length, or with a protective




case, so that they may be handled.   The tubes should be filled with




samples and standards which have been thoroughly agitated, and sufficient




time should be allowed for bubbles  to escape.




       Differences in physical design of turbidimeters will cause




differences in measured values for  turbidity even though the same




suspension is used for calibration.  To minimize such differences,




the following design criteria should be observed:




    a.  Light source - tungsten lamp operated at not less than 85% of




    rated voltage or more than rated voltage.




    b.  Distance traversed by incident light and scattered light within




    the sample tube - total not to  exceed 10 cm.




    c.  Angle of light acceptance of the detector - centered at 90° to




    the incident light path and not to exceed ± 30° from 90°.




    d.  Maximum turbidity to be measured - 40 units.






3.  Reagents




    3.1.  Turbidity-free water:  Pass distilled water through a




membrane filter having a pore size  no greater than 100 mu if such




filtered water shows a lower turbidity than the distilled water.  Discard




                                                                     SEP    1950

-------
                               -6-






the first 200 ml collected.  Otherwise, use the distilled water.




    3.2.  Stock turbidity suspension:




    a.  Solution (I):  Dissolve l.OOOg hydrazine sulfate, (NH2)2




    in distilled water and dilute to 100 ml in a volumetric flask.




    b.  Solution (II):  Dissolve lO.OOg hexamcthylenetetramine in




    distilled water and dilute to 100 ml in a volumetric flask.




    c.  In a 100 ml volumetric flask, mix 5.0 ml solution (I)  with




    5.0 ml solution (II).  Allow to stand 24 hours at 25 ± 3°  C,




    then dilute to the mark and mix.




    3.3.  Standard turbidity suspension:  Dilute 10.00 ml stock




turbidity suspension to 100 ml with turbidity free water.  The turbidity




of this suspension is defined as 40 units.




    3.4.  Dilute turbidity standards:  Dilute portions of the standard




turbidity suspension with turbidity-free water as required.




    A new stock turbidity suspension should be prepared each month.  The




standard turbidity suspension and dilute turbidity standards should be




prepared weekly by dilution of the stock turbidity suspension.






4.  Procedure:




    4.1.  Turbidimeter calibration:  The manufacturer's operating




instructions should be followed.   Measure standards on the turbidimeter




covering the range of interest.  If the instrument is already calibrated




in standard turbidity units, this procedure will check the accuracy of




the calibration scales.   At least one standard should be run in each




instrument range to be used.  Some instruments permit adjustment of




sensitivity so that scale values will correspond to turbidities.

-------
                                -7-







Rcliancc on a manufacturer's solid scattering standard for  setting




overall instrument sensitivity for all ranges is  not  an acceptable




practice unless the turbidimetcr has been shown to be free  of drift




on all ranges.  If a pre-calibrated scale is not  supplied,  then cali-




bration curves should be prepared for each range  of the instrument.




    4.2.  Turbidities less than 40 units:  Shake  the  sample to thoroughly




disperse the solids.  Wait until aj.r bubbles disappear than pour the




sample into the turbidimeter tube.  Read the turbidity directly from




the instrument scale or from the appropriate calibration curve.




    4.3.  Turbidities exceeding 40 units:  Dilute the sample with one




or more volumes of turbidity-free water until the turbidity falls below




40 units.  The turbidity of the original sample is then computed from




the turbidity of the diluted sample and the dilution factor.  For




example, if 5 volumes of turbidity-free water were added to 1 volume




of sample, and the diluted sample showed a turbidity of 30 units, then




the turbidity of the original sample was 180 units.






5.  Interpretation of Results




    5.1  Turbidity readings are reported in accordance with the




following:
Turbidity
Range ,
units
0.0-1.0
1-10
10-40
40-100
100-400
400-1000
>1000

Record
to nearest:
0.05
0.1
1
5
10
50
100
                                                                            1358

-------
                              —8—
    5.2.  For comparison of water treatment efficiencies, it may be




desirable to estimate more closely than is specified in the table:




but, because of the uncertainties and discrepancies in turbidity




measurements, it cannot be expected that two or more laboratories will




duplicate results on the same sample more closely than specified.
                                                                     SEP

-------
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet Parameter No.  to  be  assigned
Parameter    :   zinc,  Dissolved
Method       :   Filtration through. 0.45 p  membrane filter .
                Measurement by atomic absorption.
Reference    :  Method is specific to the particular instrument.
                General procedure is being selected.
Modifications:  None
                                                Date:  SePtemher 1468

-------
Parameter
             Federal Water Pollution Control Administration

                   Analytical Quality Control Branch

                       OFFICIAL INTERIM METHODS

                                  for

                   Analyses of Surface Water Samples
                                         Storet  Parameter No.  01090
Zinc, Total
Method
Atomic absorption.
Reference    :   Method  is  specific  to  the particular instrument.
                General procedure is being selected.
Modifications:   None
                                                Date:  September 1968

-------