Measurement of Purgeable Organic
          Compounds 1n Drinking Water
    by Gas Chromatography/Mass Spectrometry
                 February 1983
              Ann Alford-Stevens
             James W. Elchelberger
               W1111am L.  Budde
      Physical  and  Chemical  Methods Branch
Environmental Monitoring and Support Laboratory
       Office of  Research  and  Development
      U.S.  Environmental  Protection Agency
             Cincinnati,  Ohio  45268

-------
  o A T'E

  i
            UNITED STATES ENVIRONMENTAL PROTECTION Af.rNf.
                        FEB 2 2 1983
  r
SUBJECT Nationwide Approval of Alternate Test Procedure
                   - of Trihalomethanes
  FROM victor $/ Kimm,  Director
      -Office of Drinking Water (WH-550

    T0' Regional Administrators
       Listed  below is  an  alternate  test  procedure  for determining
       trihalometh.~r.es  by  a  gas  chromatography-mass spectrcrr.etry
       (GC-MS)  method which  I  have approved  for  nationwide use for
       "National  Interim Primary Drinking Water  Regulation" (NIPDWR)
      •compliance  monitoring.

       This  method was  developed by  the  Environmental  Monitoring and
       Support  Laboratory  (EMSL) in  Cincinnati  for  the measurement
       of  purgeable organic  compounds  in  drinking water.   Unlike the
       previously  approved Method 501.3  which  involves selected-ion
       monitoring  with  a mass  spectrometer,  this method permits the
       acquisition of complete mass  spectra.   In addition, this
       method  can  be used  for  the measurement  of many  other volatile
       organic  compounds along with  the  trihalomethanes.   EMSL's
       data  show that the  precision  and  accuracy of this  method is
       comparable  to that  of EPA's approved  methods.   (Previously
       approved methods  for  trihalomethanes  are  501.1, 501.2 and
       501.3.)  Although detection limits for  trihalomethanes are
       not as  low  as those obtained  with  other approved methodology,
       they  are entirely adequate for  the current maximum contaminant
       level.
      Measurement
      Total Trihalomethanes
                                Method

                                ^-Method  524  -  "Measurement  of
                                 Purgeable Organic  Compounds
                                 in Drinking Water  by  Gas
                                 Chromatography/Mass
                                 Spectrometry".
•'•Additional Information on this method is available  from the
 Environmental Monitoring and Support Laboratory,  26 West  St
 Street, Cincinnati, Ohio 45260.

cc :   Robert L. Booth, Acting Director, EMSL
     Ann Alford-Stevens , Chemist, EMSL
     Regional Water Supply Representatives
     Regional Quality Assurance Officers
                                                                     Clair
EPA F<"« 1320-4
           «. 3-76)

-------
                                     INDEX
Section
Number                     Subject

 1              Scope and Application
 2              Summary of Method
 3              Definitions
 4              Interferences
 5              Safety
 6              Apparatus and Equipment
 7              Reagents and Consumable Materials
 8              Sample Collection, Preservation and Handling
 9              Calibration
10              Quality Control
11              Procedure
12              Calculations
13              Precision and Accuracy
                References
                                    TABLES
1.  Ion Abundance Criteria for p-Bromofluorobenzene.
2.  Single Laboratory Method Efficiency Data for Purgeable  Organic  Compounds
    Measured with GC/MS.
3.  Acceptable Storage Times for River and Drinking Water Samples Containing
    Halogenated Aliphatic Analytes.
4.  Storage Time Data for River and Drinking Water Samples  Containing
    Aromatic Analytes.
5.  Single Laboratory Accuracy and Precision Data for Purgeable  Organic
    Compounds Measured with GC/MS.

-------
            Method  524.  Measurement of Purgeable Organic Compounds
           1n Drinking Water By Gas Chromatography/Mass Spectrometry
1.  SCOPE AND APPLICATION

    1.1  This 1s a general purpose method that provides procedures for
         Identification and measurement of purgeable organic compounds 1n
         finished drinking water, raw source water, or drinking water 1n any
         treatment stage.  The method 1s applicable to a wide range of
         organic compounds that have sufficiently high volatility and low
         water solubility to be removed from water samples with purge and
         trap procedures.  Single-laboratory method efficiency, accuracy and
         precision data have been determined for the following compounds:
                                    Chemical Abstracts Service
                                       Registry Number (CASRNT
Analyte                       Registry Numpe
benzene                             71-43-2
b1s(2-ch1oro1sopropyl) ether     39638-32-9
bromochloromethane                  74-97-5
bromodlchloromethane                75-27-4
4-bromofluorobenzene               460-00-4
bromoform                           75-25-2
carbon tetrachlorlde                56-23-5
chlorobenzene                      108-90-7
chlorodlbromomethane               124-48-1
chloroform                          67-66-3
chloromethane                       74-87-3
l,2-d1bromo-3-chloropropane         96-12-8
l,2-d1chlorobenzene                 95-50-1
l,3-d1chlorobenzene                541-73-1
l,2-d1chloroethane                 107-06-2
I,l-d1chloroethene                  75-35-4
trans-l,2-d1chloroethene           156-60-5
c1s-l,3-d1chloropropene          10061-01-5
methylene chloride                  75-09-2
styrene (ethenylbenzene)           TOO-42-5
1,1,2,2-tetrachloroethane           79-34-5
tetrachloroethene                  127-18-4
toluene                            108-88-3
1,1,1-tHchloroethane               71-55-6
1,1,2-trlchloroethane               79-00-5
tHchloroethene                     79-01-6
vinyl chloride                      75-01-4
p-xylene*                           106-42-3
STORET
Number
 34030
 34283
 77297
 32101

 32104
 32102
 34301
 32105
 32106
 34418

 34536
 34566
 34531
 34501
 34546
 34699
 34423
 77128
 34516
 34475
 34010
 34506
 34511
 39180
 30175
         A laboratory may use this method to detect and measure additional
         analytes after the laboratory obtains acceptable (defined 1n
         Section 10) accuracy and precision data for each added analyte.

-------
    1.2  Detection Limits  (MOLs)  (1) are compound dependent, varying with
         purging efficiency and concentration, where MDL  is defined as the
         statistically calculated minimum amount that can be measured with
         99* confidence that the  reported value is greater than zero.  For
         the listed analytes in reagent water, MOLs vary from 0.07 to 11.2
         ug/L.  The applicable concentration range of this method is
         compound and instrument  dependent but is approximately 0.2 yg to
         200 ug of analyte per liter of undiluted sample.  Analytes that are
         inefficiently purged from water will not be detected when present
         at low concentrations, but they can be measured with acceptable
         accuracy and precision when present in sufficient amounts.

    1.3  Determination of some individual components of complex mixtures may
         be hampered by Insufficient chromatographic resolution and/or by
         large differences In concentrations of individual components.

2.  SUMMARY OF METHOD

    Highly volatile organic compounds with low water solubility are removed
    (purged) from the sample matrix by bubbling helium through a 25-mL
    aqueous sample.  Purged sample components are trapped in a stainless
    steel tube containing suitable sorbent materials.  When purging 1s
    complete, the sorbent tube 1s heated and backflushed with helium to
    desorb purged sample components Into a gas chromatograph (GC) Interfaced
    to a mass spectrometer (MS).  Compounds eluting from the GC column are
    tentatively Identified by comparing their mass spectra to reference
    spectra in a data base.  Tentative Identifications are confirmed by
    analyzing standards under the same conditions used for samples and
    comparing resultant mass spectra and GC retention times.  Each Identi-
    fied component 1s measured by relating the MS response for an appro-
    priate selected ion produced  by that compound to the MS response for the
    same ion from that same compound in an external standard or for another
    ion produced by a compound that 1s used as an Internal standard.

3.  DEFINITIONS

    External standard — a known  amount of a pure compound that is analyzed
    with the same procedures and  conditions that are used to analyze samples
    containing that compound.  From measured detector responses to known
    amounts of the external standards a concentration of that same compound
    can be calculated from measured detector response to that compound in a
    sample analyzed with the same procedures.

    Internal standard — a pure compound added to a sample in known amounts
    and used to calibrate concentration measurements of other compounds that
    are sample components.  The Internal standard must be a compound that is
    not a sample component.

    Field duplicates — two samples taken at the same time and place under
    identical circumstances and treated exactly the same throughout field
    and laboratory procedures.  Analysis of field duplicates Indicates the
    precision associated with sample collection, preservation and storage,
    as well as with laboratory procedures.

-------
Field reagent blank — reagent water placed  in a sample container  1n  the
laboratory and treated as a sample in all respects,  including exposure
to sampling, site conditions, storage, preservation and all analytical
procedures.

Laboratory control standard — a solution of analytes prepared  in  the
laboratory by dissolving known amounts of pure compounds in a known
amount of reagent water.  In this method, the laboratory control
standard is prepared by adding appropriate volumes of the secondary
dilution standard solution and the internal standard/surrogate  compound
spiking solution to reagent water.

Laboratory duplicates — two aliquots of the same sample that are
treated exactly the same throughout laboratory analytical procedures.
Analysis of laboratory duplicates Indicates precision associated with
laboratory procedures but not with sample collection, preservation or
storage procedures.

Laboratory reagent blank — a 25-mL portion of reagent water placed in
the purging apparatus and analyzed as if it were a sample.

Performance evaluation sample — a methanol solution of method  analytes
distributed by the Quality Assurance Branch  (QAB), Environmental
Monitoring and Support Laboratory, USEPA, Cincinnati, Ohio, to  multiple
laboratories for analysis.  A small volume of the methanol solution 1s
added to a known volume of reagent water and analyzed with procedures
used for samples.  Results of analyses are used by the QAB to determine
statistically the accuracy and precision that can be expected when a
method 1s performed by competent analysts.  Analyte  true values are
unknown to the analyst.

Quality control check sample — a methanol solution  containing  known
concentrations of analytes prepared by a laboratory  other than  the
laboratory performing the analysis.  The analyzing laboratory uses this
solution to demonstrate that it can obtain acceptable Identifications
and measurements with a method.  A small volume of the methanol solution
1s added to a known volume of reagent water and analyzed with procedures
used for samples.  True values of analytes are known by the analyst.

Secondary dilution standard — a methanol solution of analytes  prepared
in the laboratory from stock standard solutions and  diluted as  needed to
prepare aqueous calibration solutions and laboratory control standards.

Stock standard solution — a concentrated solution containing a
certified standard that 1s a method analyte, or a concentrated  methanol
solution of an-analyte prepared in the laboratory with an assayed
reference compound.  Stock standard solutions are used to prepare
secondary standard solutions.

Surrogate compound — a compound that 1s not expected to be found  in  the
sample, 1s added to the original environmental sample to monitor perform-
ance, and is measured with the same procedures used  to measure  sample
components.

-------
4.  INTERFERENCES

    4.1  Samples may be contaminated during shipment or storage by diffusion
         of volatile organics through the sample bottle septum seal.  Field
         reagent blanks must be analyzed to determine when sampling and
         storage procedures have not prevented contamination.

    4.2  During analysis, major contaminant sources are volatile materials
         in the laboratory and impurities in the inert purging gas and in
         the sorbent trap.  Analyses of field reagent blanks and laboratory
         reagent blanks provide information about the presence of contami-
         nants.

    4.3  Interfering contamination may occur when a sample containing low
         concentrations of volatile organic compounds is analyzed
         immediately after a sample containing relatively high concentra-
         tions of volatile organic compounds.  A preventive technique is
         between-sample rinsing of the purging apparatus and sample syringes
         with two portions of reagent water.  After analysis of a sample
         containing high concentrations of volatile organic compounds, the
         system should be baked for 10 min by passing helium through the
         sample purging chamber into the heated (180°) sorbent trap.  One
         or more laboratory reagent blanks should be analyzed to ensure that
         accurate values are obtained for the next sample.

5.  SAFETY

    5.1  The toxldty or cardnogenlcity of chemicals used in this method
         has not been precisely defined; each chemical should be treated as
         a potential health hazard, and exposure to these chemicals should
         be minimized.  Each laboratory 1s responsible for maintaining
         awareness of OSHA regulations regarding safe handling of chemicals
         used in this method.  Additional references to laboratory safety
         are cited (2-4).

    5.2  The following method analytes have been tentatively classified as
         known or suspected human or mammalian carcinogens:  benzene, carbon
         tetrachloride, chloroform, and vinyl chloride,  tore standard
         materials and stock standard solutions of these compounds should be
         handled in a hood.

6.  APPARATUS AND EQUIPMENT

    6.1  Sample containers — 120-ml or larger glass bottles each equipped
         with a sCrew cap and a polytetrafluoroethylene-faced s111 cone
         septum.
    6.2  dirge and trap device consisting of sample purging chamber, sorbent
         trap and desorber.  (Acceptable devices are commercially available.)

         6.2.1  The all glass sample purging chamber holds 25-mL samples
                with < 15 ml of gaseous headspace between the water column
                and the trap.  The helium purge gas passes through the water
                column as finely divided bubbles (optimum diameter of <3 mm

-------
               at the origin).  The purge  gas must  be  Introduced  at  a
               point <5 mm from the base of  the  water  column.

     6.2.2  The stainless steel sorbent trap 1s  25  cm  long  by  2.5 mm ID
            and 1s packed with 8 cm of Tenax-GCR, 8 cm of silica  gel,
            and 8 cm of charcoal, 1n that  order  with respect to the
            Inlet end of the trap.  The charcoal  1s not  necessary for
            listed analytes but 1s necessary if  fluorine-substituted
            methanes and ethanes (fluorocarbons)  are among  additional
            analytes.  When analytes do not  Include fluorocarbons, the
            charcoal may be eliminated, and the  amount of Tenax-GCR
            may be Increased.  A trap with different dimensions can  be
            used 1f 1t has been evaluated  and found to perform satisfac-
            torily (I.e., provides method  efficiencies equal to or
            better than those 1n Table 2).   Before  Initial  use, the  trap
            should be conditioned overnight  at 180°C by backflushlng
            with helium flow of at least 20 mL/m1n.  Each day  the trap
            should be conditioned for 10 m1n at  180°C  with  back-
            flushing.

     6.2.3  The desorber should be capable of rapidly  heating  the trap
            to 180°C.  The trap section containing  Tenax-GCR should
            not be heated to higher than 180°C,  and the temperature  of
            the other sections should not  exceed 200°C.

6.3  SYRINGES AND SYRINGE VALVES

     6.3.1  Two 25-nt glass hypodermic syringes  with Luer-LokR tip (1f
            applicable to the purging device being  used).

     6.3.2  One 5-mL gas-tight syringe with  shutoff valve.

     6.3.3  Two two-way syringe valves with  Luer ends  (1f applicable to
            the purging device being used).

     6.3.4  Micro syringes, various sizes.

6.4  MISCELLANEOUS

     6.4.1  Standard solution storage containers  — 10-mL bottles with
            polytetrafluoroethylene-lined  screw  caps.

     6.4.2  Analytical balance capable of  weighing  0.0001 g accurately.

     6.4.3  Helium purge gas, as contaminant free as possible.

6.5  GAS CHROMATOGRAPH/MASS SPECTROMETER/DATA SYSTEM (GC/MS/OS)

     6.5.1  The GC must be capable of temperature programming. Any
            column (either packed or capillary)  that provides  data with
            adequate accuracy and precision  (Sect.  10) can  be  used.   If
            a packed column 1s used, the GC usually 1s Interfaced to the
            MS with an all-glass enrichment device  and an all-glass

-------
                transfer line, but any enrichment device or transfer  line
                can be used if performance specifications described 1n this
                method can be demonstrated with 1t.  If a capillary column
                1s used, an enrichment device 1s not needed.  A recommended
                packed GC column for the listed analytes 1s 1.8 m long by 2
                mm ID glass packed with IX SP-1000 on 60/80 mesh Carbopack
                B.  Recommended operating parameters for that column  are:
                helium carrier gas flow rate of 30 mL/m1n and temperature of
                45°C for 4 m1n, Increased to 230°C at a rate of
                8°C/m1n, and Isothermal at 230°C for at least 25 m1n  or
                until all expected analytes elute.  An alternative recom-
                mended packed column 1s 1.8 m long by 2 mm ID glass or
                stainless steel packed with 0.2X Carbowax 1500 on 80/100
                mesh Carbopack C.

         6.5.2  Mass spectral data are obtained with electron-Impact
                1on1zat1on at a nominal electron energy of 70 eV.  The mass
                spectrometer must be capable of scanning from 35 to 450 amu
                every 7 s or less and must produce a mass spectrum that
                meets all criteria 1n Table 1 when 50 ng or less of
                p_-bromof1uorobenzene (BFB) 1s Introduced Into the GC.  To
                ensure sufficient precision of mass spectral data, the
                desirable MS scan rate allows acquisition of at least five
                spectra while a sample component elutes from the GC.  With
                capillary columns which produce narrower peaks than packed
                columns that criterion may not be feasible and adequate
                precision with fewer spectra per GC peak must be
                demonstrated (Sect. 10).

         6.5.3  An Interfaced data system (OS) 1s required to acquire,
                store, reduce and output mass spectral data.  The computer
                software must allow searching any GC/MS data file for Ions
                of a specific mass and plotting 1on abundances versus time
                or scan number.  This type of plot 1s defined as an
                extracted 1on current profile (EICP).  Software must  also
                allow Integrating the abundance 1n any EICP between speci-
                fied time or scan number limits.

7.   REAGENTS AND CONSUMABLE MATERIALS

    7.1  SORBENT TRAP PACKING MATERIALS

         7.1.1  Polymer based on 2,6-d1pheny1-p_-pheny1ene oxide — 60/80
                mesh Tenax-GC", chromatograpMc grade, or equivalent.

         7.1.2  Coconut charcoal — 26 mesh.

         7.1.3  Silica gel — 35/60 mesh, Davison Chemical grade 15,  or
                equivalent.

    7.2  REAGENTS

         7.2.1  Methanol — pesticide quality or equivalent.

-------
     7.2.2  Reagent water — water 1n which an Interferent 1s not
            observed at the method detection limit of the compound of
            Interest. Prepare reagent water by. passing tap water through
            a filter bed containing about 0.5 kg of activated carbon, by
            using a water purification system, or by boiling distilled
            water for 15 min followed by a 1 h purge with Inert gas
            while the water temperature 1s held at 90°C.  Store 1n
            clean, narrow-mouth bottles with polytetrafluoroethylene-
            Uned septa and screw caps.

     7.2.3  Sodium tMosulfate or sodium sulflte ~ granular, ACS
            reagent grade.

7.3  STOCK STANDARD SOLUTIONS — These solutions may be purchased as
     certified solutions or prepared from pure standard materials using
     the following procedures:

     7.3.1  Place about 9.8 mL of methanol 1n a 10-mL ground-glass
            stoppered volumetric flask.  Allow the flask to stand
            unstoppered for about 10 min or until all alcohol-wetted
            surfaces have dried.  Weigh the flask to the nearest 0.1
            mg.

     7.3.2  If the analyte 1s a liquid at room temperature, with a
            100-uL syringe Immediately add two or more drops of assayed
            reference compound to the flask.  (The liquid must fall
            directly Into the alcohol without contacting the flask).  If
            the analyte 1s a gas at room temperature, fill a 5-mL valved
            gas-tight syringe with the standard to the 5.0-mL mark,
            lower the needle to 5 mm above the methanol meniscus, and
            slowly Inject the standard Into the neck of the flask.  (The
            gas will rapidly dissolve 1n the methanol.)

     7.3.3  Reweigh the flask, dilute to volume, stopper, and mix by
            Inverting several times.

     7.3.4  From the net weight gain, calculate the concentration in
            micrograms per micro liter.  When assayed compound purity 1s
            >96X, the unconnected weight may be used to calculate
            concentration.

     7.3.5  Store stock standard solutions with minimal headspace in
            polytetrafluoroethylene-Hned screw-capped bottles.
            Methanol solutions of listed liquid analytes are stable for
            at least four weeks when stored at 4<>C.  Methanol
            solutions prepared from listed gaseous analytes are not
            stable for more than one week when stored at <0°C; at room
            temperature, they must be discarded after one day.

7.4  SECONDARY DILUTION STANDARD — Use stock standard solutions to
     prepare a secondary dilution standard solution that contains the
     analytes 1n methanol.  The secondary dilution standard should be
     prepared at a concentration that can be easily diluted to prepare

-------
         aqueous calibration solutions (Section 9.2.1.2) at concentrations
         that will bracket the working concentration range.  Store the
         secondary dilution standard solution with minimal headspace and
         check frequently for signs of deterioration or evaporation,
         especially just before preparing calibration solutions from 1t.

    7.5  INTERNAL STANDARD SPIKING SOLUTION AND SURROGATE COMPOUND SPIKING
         SOLUTION — Prepare a solution of fluorobenzene in methanol at a
         concentration that allows use of 2 to 10 uL to add an appropriate
         amount of fluorobenzene to each sample; this amount should be
         approximately the same as the amount of the analyte to be measured.
         If the Internal standard technique 1s used, fluorobenzene serves as
         the Internal standard.  If the external standard technique is used,
         fluorobenzene is a surrogate compound added to each sample to
         monitor method performance.  Fluorobenzene was selected because it
         is stable in aqueous solutions, is. efficiently purged, does not
         occur naturally, and 1s not commercially produced in bulk
         quantities but 1s available as a laboratory reagent chemical.

8.  SAMPLE COLLECTION, PRESERVATION AND HANDLING

    8.1  Collect all samples in duplicate.  Fill sample bottles to over-
         flowing.  No air bubbles should pass through the sample as the
         bottle is filled, or be trapped in the sample when the bottle 1s
         sealed.  Keep samples sealed from collection time until analysis.
         Maximum storage times vary with analytes of concern.  Recent
         studies (5-6) provided data indicating appropriate storage times
         for samples (river and drinking water) containing compounds that
         are potential method analytes [Tables 3 and 4).

         8.1.1  When sampling from a water tap, open the tap and allow the
                system to flush until the water temperature has stabilized
                (usually about 10 m1n).  Adjust the flow to about 500 mL/m1n
                and collect duplicate samples from the flowing stream.

         8.1.2  When sampling from an open body of water, fill a 1-qt
                wide-mouth bottle with sample from a representative area,
                and carefully fill duplicate sample bottles from the 1-qt
                bottle.

    8.2  SAMPLE PRESERVATION

         8.2.1  If styrene (which reacts with chlorine) and/or trlhalo-
                methanes are to be measured in a sample expected to contain
                residual chlorine, add a reducing agent, sodium thlosulfate
                or sodium sulfite (30 mg per 120-mL sample for up to 5 ppm
                chlorine) to the empty sample bottle before it 1s shipped to
                the sampling site.

                NOTE:  Some possible analytes may be unstable in the
                presence of reducing agent.  Data (5) Indicate that sodium
                sulfite should not be used if analytes Include
                chloromethane, I,l-d1chloroethene, I,l-d1chloropropene,

-------
            2-chloroethyl ethyl ether, or 1,1,2,2-tetrachloroethane;
            sodium thlosulfate 1s not recommended 1f analytes Include
            chloromethane or 1,2-d1bromoethane.

     8.2.2  Much remains to be learned about biological degradation of
            aromatic hydrocarbon analytes.  Currently, two preservation
            techniques are recommended but both have negative aspects.

            8.2.2.1 Hydrochloric add may be used at the sampling site
                    to adjust the sample pH to < 2; the major disadvan-
                    tages of this procedure are that shipping add 1s
                    restricted by federal regulations and that effects
                    of low pH on other analytes (such as organohalldes)
                    are largely unknown.

            8.2.2.2 Mercuric chloride may be added to the sample bottle
                    1n amounts to produce a concentration of 10 mg/L.
                    This may be added to the sample at the sampling site
                    or to the sample bottle 1n the laboratory before
                    shipping to the sampling site.  A major disadvantage
                    of mercuric chloride 1s that 1t 1s a highly toxic
                    chemical; 1t must be handled with caution, and
                    samples containing 1t must be disposed with appro-
                    priate procedures.

            8.2.2.3 If analytes Include both aromatic hydrocarbons and
                    styrene or trlhalomethanes, current recommendations
                    are either to add both preservative types (reducing
                    agent along with add or mercuric chloride) or to
                    collect two samples with the appropriate preserva-
                    tive type 1n each.

     8.2.3  After addition of preservatlve(s), seal the sample bottle
            and shake vigorously for 1 mln.

8.3  FIELD BLANKS

     8.3.1  Duplicate field reagent blanks must be handled along with
            each sample set, which 1s composed of the samples collected
            from the same general sample site at approximately the same
            time.  At the laboratory, fill field blank sample bottles
            with reagent water, seal, and ship to the sampling site
            along with empty sample bottles and back to the laboratory
            with filled sample bottles.  Wherever a set of samples 1s
            shipped and stored, 1t 1s accompanied by appropriate blanks.

     8.3.2  When reducing agent or preservative(s) 1s added to samples,
            use the same procedures used for samples to add the same
            amount to blanks.  The reducing agent can be added 1n the
            laboratory.

-------
9.  CALIBRATION

    9.1  INITIAL CALIBRATION

         9.1.1  CALIBRATION SOLUTIONS
                9.1.1.1 At least three calibration solutions, each contain-
                        ing a standard of each analyte, are needed.  (More
                        than three calibration solutions may be required if
                        analytes are numerous.)  One calibration solution
                        should contain each analyte at a concentration
                        approaching but greater than the method detection
                        limit (Table 5) for that compound; the other two
                        solutions should contain analytes at concentrations
                        that bracket the range expected in samples.  For
                        example, if the detection limit for a particular
                        analyte is 0.2 ug/L, and a 25-mL sample expected to
                        contain approximately 5 ug/L 1s analyzed, aqueous
                        solutions of standards should be prepared at
                        concentrations of 0.3 ug/L, 5 ug/L, and 10 ug/L.

                9.1.1.2 To prepare calibration solutions, add appropriate
                        volumes (uL) of the secondary dilution standard
                        solution to aliquots of reagent water.  Remove the
                        plunger from a 25-mL syringe and attach a closed
                        syringe valve.  Fill the syringe with reagent water,
                        replace the plunger, and compress the water.  Open
                        the syringe valve and vent air.  Adjust the reagent
                        water volume to 25 nt and add a carefully measured
                        aliquot of 2.0 to 18.0 uL of the secondary dilution
                        standard through the valve bore.  Add the appro-
                        priate amount (>_ 2.0 uL) of the Internal standard
                        surrogate spiking solution through the valve bore,
                        but do not add more than 20 uL total volume of
                        methanol solution.

                        NOTE:  If appropriate concentrations cannot be
                        prepared without adding more than 20 uL of the
                        secondary dilution standard to 25 mL of reagent
                        water, prepare a new secondary dilution standard.
                        If less than 2.0 uL must be added to obtain appro-
                        priate concentrations, prepare a larger volume of
                        the calibration solution 1n a volumetric flask.  Mix
                        by Inverting the flask several times, and transfer a
                        25-mL portion Into the sample syringe.  The remain-
                        Ing solution may be stored in screwcap vials with no
                        headspace.  If aromatic compounds are among calibra-
                        tion solution components, do not store for more than

-------
                     1 h.  Other aqueous solutions can be  stored  for  up
                     to 24 h.

     9.1.2  Analyze  triplicate aliquots of each calibration  solution
            with the procedures to be used to analyze samples.

            9.1.2.1  If the external standard technique  1s being  used,
                    prepare a concentration calibration curve for each
                     analyte by plotting Integrated abundances of the  1on
                    characteristic of that compound as  a  function of  the
                    concentration.  If the ratio of 1on abundance to
                     amount of analyte 1s constant (< 10X  relative
                    standard deviation) throughout the  concentration
                     range, the average ratio may be used  Instead of  a
                     calibration curve.

            9.1.2.2  If the Internal standard technique  1s being  used,
                    calculate the mass spectrometer response to  each
                     compound relative to fluorobenzene, the  Internal
                     standard.  Calculate the response factor (RF) with
                     the equation,
-  V Q
                       RF
            where   Ax *   Integrated abundance of the selected  1on  for
                           the analyte standard;
                    As *   Integrated abundance of the selected  1on  for
                           the Internal standard;
                    Qs *   quantity of Internal standard;  and
                    Qx *   quantity of analyte standard.

            RF 1s a unltless number; units used to express quantities  of
            analyte and Internal standard must be equivalent.   Ideally,
            the response factor for each analyte should be Independent
            of analyte quantity for the working range of the
            calibration, but required linearity will vary  with  required
            accuracy of analyte concentration measurements.  Generally,
            acceptable variations of mean RF values are ±  15X over a
            concentration range of two orders of magnitude and  ± 10% RSD
            of values obtained from analyses of triplicate aliquots  of
            each concentration calibration solution.  For  an analyte
            with non-linear RF, a calibration curve of Areax/Areas
            plotted versus Qx may be used to determine an  analyte
            concentration.

9.2  DAILY CALIBRATION — Check calibration data each day  by measurement
     of one or more laboratory control standards or calibration  solu-
     tions.  If the expected 1on abundance was observed (Sect.  10.5) for
     50 ng of the MS performance standard but the absolute 1on  abundance

-------
         measured for any analyte varies from expected abundance by more
         than 15X, prepare and analyze a fresh calibration solution to
         determine 1f the problem 1s being caused by deterioration of the
         calibration solution or by a malfunction 1n the purge and trap
         apparatus.  When the Internal standard technique Is being used,
         verify each day that response factors have not changed.  When
         changes occur (> 10X relative standard deviation), prepare and
         analyze new standard solutions to determine new response factors.

         NOTE:  Some analysts have observed marked deterioration of MS
         response after the Initial purge and trap analysis each day; 1f
         this phenomenon is observed, perform one purge/desorb cycle before
         checking MS performance and calibration data.

10.  QUALITY CONTROL

    10.1 Minimum quality control requirements consist of:

         10.1.1 initial demonstration of laboratory analytical capability
                (efficiency, accuracy and precision procedures, Sect. 10),

         10.1.2 analysis of an MS performance standard and a laboratory
                control standard near the beginning of each 8-h work period,

         10.1.3 analysis of a field reagent blank along with each sample set,

         10.1.4 analysis of a laboratory reagent blank when the field
                reagent blank contains analytes at concentrations above the
                method detection limits,

         10.1.5 quarterly analysis of a quality control check sample, (1f
                available for analytes of concern), and

         10.1.6 continued maintenance of performance records to define the
                quality of generated data.

    10.2 METHOD EFFICIENCY — For each analyte, calculate method efficiency
         by comparing the detector response when the compound 1s introduced
         by syringe injection with the detector response when the same
         amount 1s Introduced by purging, trapping, and desorptlon.  Because
         of the calibration technique used In this method, high efficiency
         1s not required for acceptable precision and accuracy, but low
         method efficiency may cause unacceptably high detection limits.
         Measure method efficiency for each analyte whenever the analytical
         system undergoes major modification, such as replacement of trap
         packing."

         10.2.1 Analyze at least five laboratory control standards with the
                purge, trap, desorptlon and GC/MS detection procedures.
                Interspersed among these five or more analyses, inject two
                or more allquots of the secondary dilution standard solution

-------
            (Section 7.4) directly Into the SC to  introduce  each  analyte
            1n an amount equal to that Introduced  by purge and  trap
            procedures.  Use the same MS data acquisition parameters  for
            injected analytes as those used for purge and trap  proce-
            dures.

     10.2.2 Calculate the method efficiency (E) for each analyte  in each
            aliquot of the laboratory control standard with  the equation:

                                       100
                                1

            where Ap «   ion abundance of compound Introduced with purge
                         and trap techniques, and
                  A.J •   1on abundance produced by an equal  amount of
                         the same compound when Injected.

            For this calculation, use data obtained from an  Injection
            either closely preceding or following  the purge  and trap
            analysis from which data are used.

     10.2.3 Calculate the mean method efficiency for each analyte.
            Acceptable detection limits usually can be achieved even  1f
            the mean method efficiency 1s only 20  to 30X.

10.3 ACCURACY — To determine accuracy, analyze duplicate allquots of a
     quality control (QC) check sample containing  known amounts of
     analytes of concern.  QC check samples for some, but not all  listed
     analytes, currently are available from the U.S. Environmental
     Protection Agency, Environmental Monitoring and Support Laboratory,
     Quality Assurance Branch, Cincinnati, Ohio  45268; alternatively
     certified standard solutions may be purchased from commercial
     vendors.

     10.3.1 When using the external standard procedure, calculate
            accuracy as the ion abundance found 1n the QC sample
            solution expressed as a percentage (P) of the ion abundance
            found in the external standard solution:
                                         100
            where Ax »   abundance of 1on used to measure  an  analyte  in
                         an aliquot of the QC check  sample, and
                  As «   abundance of 1on used to measure  an  equal
                         amount of the same analyte  treated as an
                         external standard.

     10.3.2 When using the Internal standard procedure, fluorobenzene  in
            the solution of analyte standards 1s the Internal standard.

-------
            Calculate response factors (Sect. 9.3.4) for each analyte
            relative to fluorobenzene. With these response factors,
            calculate accuracy with data acquired for the QC check
            sample:

                      P -   Ax  .  100
                            f(3  .   RF


            where Ax • abundance of 1on used to measure an analyte in
                       an aliquot of the QC check sample,
                  As • abundance of 1on used to measure
                       fluorobenzene 1n the same aliquot, and
                  RF « response factor of the particular
                       analyte relative to fluorobenzene.

               NOTE:  The Internal standard concentration 1s constant 1n
               calibration solutions and all samples for which the
               calibration solutions are used (Section 9.3).

     10.3.3 For each analyte, the mean accuracy should be 1n the range
            of 85 to 115t.  For some listed analytes, this may not be
            feasible for low concentration measurements.

10.4 PRECISION

     10.4.1 For each analyte, calculate method precision as the standard
            deviation (s, 1n ug/l) of the replicate measured values
            obtained 1n the accuracy calculations:
   n

n
                                     1-1   X12  "
                                            n (n-1)

            where n • number of measurements for each analyte, and
                  X » Individual measured value.

     10.4.2 For the set of measured values for each analyte, calculate
            the dispersion as the percent relative standard deviation
            (RSD):

                          RSO •  _s_  .  100
                                  «•
                                  C

            where s * standard deviation, and
                  C » mean observed concentration.

     10.4.3 Adequate precision 1s obtained when the relative standard
            deviation 1s <. 20X.  For some listed analytes, this may not
            be feasible for low concentration measurements.

-------
10.5 MS PERFORMANCE STANDARD

     10.5.1 Near the beginning of each 8-h work period 1n which analytes
            are to be measured, measure the mass spectrum produced by 50
            ng of p_-bromof1uorobenzene (BFB) to ensure that 1t meets
            performance criteria (Table 1).  BFB may be Introduced Into
            the MS either by syringe Injection or through the purge and
            trap system.  It may be a component of the laboratory
            control standard analyzed dally to check calibration (Sect.
            7).  Measure the entire mass spectrum at an MS scan rate
            that produces at least five spectra for the BFB GC peak but
            does not exceed 7 s per spectrum.  Although acquisition of
            five spectra per BFB GC peak may not be feasible when capil-
            lary columns are.used, BFB performance criteria still must
            be met.  If the BFB spectrum 1s unacceptable, adjust GC/MS
            operating parameters until an acceptable spectrum 1s pro-
            duced before samples are analyzed.

     10.5.2 Record the absolute 1on abundance detected for 50 ng of
            BFB.  If ion abundance varies more than ± 10% from the
            expected number, check the GC/MS system to locate and
            correct the problem.  Preparation of a new calibration curve
            may be necessary if the system 1s operating acceptably but
            with decreased sensitivity.

10.6 LABORATORY CONTROL STANDARD — To demonstrate that the calibration •
     curve 1s still valid, analyze a laboratory control standard at the
     beginning of each 8-h work period.

     10.6.1 For each analyte to be measured, select a concentration
            representative of Its occurrence 1n drinking water samples.

     10.6.2 Prepare the laboratory control standard with either of the
            following procedures:

            10.6.2.1  From stock standard solutions, prepare a
                      laboratory control standard concentrate in
                      methanol.  This solution should contain analytes
                      at concentrations 2500 times those selected as
                      representative concentrations.  Add 10 uL of the
                      laboratory control standard concentrate to a 25-mL
                      aliquot of reagent water.

            10.6.2.2  Add 2 to 18 uL of the secondary dilution standard
                      to 25 mL of reagent water contained 1n the sample
                      syringe.

     10.6.3 Add an appropriate volume of the Internal standard/surrogate
            spiking solution and analyze with the same procedures (Sect.
            11) to be used for samples.

-------
      10.6.4 Determine calibration acceptability and appropriate remedial
             actions, if needed.  (For the external standard technique,
             see Sect. .9.1.2.1; for the internal standard technique, see
             Sect. 9.1.2.2.)

10.7  MONITORING THE SURROGATE COMPOUND/INTERNAL STANDARD - Because all
      samples and laboratory control standards contain equal amounts of
      the internal standard/surrogate compound, use the absolute ion
      abundance for the characteristic ion of that compound, fluoro-
      benzene, to monitor system performance.  If for any sample, the
      absolute ion abundance varies more than 15X from that observed in
      the previous sample or laboratory control standard, do not report
      analyte values obtained for that sample, and take remedial actions
      to solve the system performance problem.

10.8  FIELD REAGENT BLANKS — Analyze a field reagent blank along with
      each sample set.  If a field reagent blank contains analytes at
      concentrations above the method detection limits, analyze a
      laboratory reagent blank.  If one or more analytes that are not
      detected at concentrations above method detection limits in the
      laboratory reagent blank are detected in significant amounts in the
      field blank, sampling or storage procedures have not prevented
      sample contamination, and the appropriate analyte measurement(s)
      must be discarded.

10.9  At least quarterly, analyze a quality control check sample obtained
      from the U.S. Environmental Protection Agency, Environmental
      Monitoring and Support Laboratory, Quality Assurance Branch,
      Cincinnati, Ohio.  Quality control check samples currently are
      available for some but not all listed analytes.  If measured
      analyte concentrations are not within ±20X of true values, check
      the entire analytical procedure to locate and correct the problem
      source.

10.10 Additional QC procedures may be necessary, depending on the purpose
      of the analysis performed with this method.

      10.10.1   Laboratory Duplicates — To determine precision
                associated with laboratory techniques, analyze two
                aliquots (Sect. 11.1.2) of a sample in which some
                analytes were detected in measurable quantities.
                Calculate the range (R) of concentrations measured for
                each duplicate pair:

                     R • Ci - Cfc,

                where  C| represents the larger and,
                       Cg represents the smaller of the two
                       measurements.

                Calculate percent relative range (RR) of duplicate
                analyses using the formula:

-------
                        RR
              R
             ^•^•i

              C
100
         10.10.2
         10.10.3
where  R * range of concentrations measured, and
       C a mean concentration measured.
Generally, 1f RR 1s greater than 30X, precision 1s
Inadequate, and laboratory techniques must be Improved.

Field Duplicates — Analyze 10X of samples in which some
analytes were detected in measurable quantities to
Indicate precision limitations Imposed by sampling,
transport and storage techniques as well as laboratory
techniques.  If acceptable results are obtained from
analysis of field duplicates, analysis of laboratory
duplicates 1s usually not necessary.

Matrix Effects Determination — To Indicate matrix
effects on method efficiency, accuracy and precision when
raw source waters or drinking water during treatment 1s
to be analyzed, analyze allquots to which known amounts
of analytes have been added.  Because analytes may be
present 1n the unspiked allquots, analysis of one or more
unsplked allquots 1s necessary to determine Initial
concentrations, which are then subtracted from concentra-
tions measured 1n spiked allquots.  For each analyte the
amount added to determine matrix effects should exceed
twice the amount measured in unsplked allquots.
11.  PROCEDURE
    11.1 ANALYSIS PROCEDURES

         11.1.1 Initial conditions — Adjust the helium purge gas flow rate
                to 40 ± 3 mL/m1n.  Attach the sorbent trap to the purging
                device, and set the device to the purge mode.  Open the
                syringe valve located on the sample Introduction needle of
                the purging chamber.

         11.1.2 Sample Introduction and purging — Remove the plunger from a
                25-mL syringe and attach a closed syringe valve.  Open the
                sample or standard bottle, which has been allowed to come to
                ambient temperature, and pour the sample Into the syringe
                barrel to just short of overflowing.  Replace the syringe
                pJunger and compress the sample.  Open the syringe valve and
                vent any residual air while adjusting the sample volume to
                25.0 ml.  (Because this process of taking an aliquot Impairs
                the Integrity of the remaining sample, a second syringe
                should be filled at the same time, 1n case a second analysis
                1s required.)  Add 2 to 20 uL of the spiking solution (Sect.
                7.5) of fluorobenzene 1n methanol through the syringe valve
                and close the valve.  Attach the syringe and Its valve

-------
            assembly to the syringe valve on the purging device.  Open
            the syringe valves and slowly inject the sample into the
            purging chamber.  Close both valves and purge the sample for
            11.0 ± 0.1 min at ambient temperature.  Because temperature
            affects purging efficiencies of some analytes, purging
            chamber temperature must be controlled to maintain constant
            (approximately ± 2°C) temperature throughout calibration
            and sample analyses.  If laboratory temperature is not
            controlled adequately, the purging chamber can be placed in
            a thermostatically controlled water bath.

     11.1.3 Oesorption and data acquisition — At the conclusion of
            purging, adjust the purge and trap apparatus to the desorb
            mode, and initiate GC temperature programming, trap heating,
            and MS data acquisition.  Oesorb for 4 min.  Transfer
            trapped sample components into the GC column by heating the
            trap to 18QOC rapidly while it is backflushed with helium
            flowing at 20 to 60 mL/min.  (If the trap cannot be heated
            rapidly, use the GC column as a secondary trap by cooling
            the column to < 30°C during desorption.)

     11.1.4 Sample chamber rinsing — During or after desorption empty
            the purging chamber with the sample introduction syringe,
            and rinse the chamber with two 25-mL portions of reagent
            water.

     11.1.5 Trap reconditioning — After desorblng the sample for 4 min,
            reset the purging device to the purge mode.  After 15 s,
            close the syringe valve on the purging device to begin gas
            flow through the trap.  After approximately 7 m1n, turn off
            the trap heater and open the syringe valve to stop gas flow
            through the trap.  When cool (<25°C), the trap 1s ready
            for the next sample.

     11.1.6 Termination of data acquisition — When sample components
            have eluted from the GC, terminate MS data acquisition and
            store data files on the data system storage device.  Use
            appropriate data output software to display full range mass
            spectra and appropriate extracted ion current profiles
            (EICPs).  If any ion abundance exceeds the system working
            range, dilute the sample aliquot in the second syringe with
            reagent water and analyze the diluted aliquot.

11.2 IDENTIFICATION PROCEDURES CRITERIA — Tentatively Identify a sample
     component by comparison of Us mass spectrum (after background
     subtraction) to a reference spectrum in a collection.  Use the
     following criteria to confirm a tentative identification:

     11.2.1 The GC retention time of the sample component must be within
            £ s of the time observed for that same compound when a
            calibration solution was analyzed.  Calculate the value of t_
            with the equation:

-------
                        t-(RT)V3

                where RT » observed retention time (1n seconds) of the
                compound when a calibration solution was analyzed.

         11.2.2 All ions that are present above 1W relative abundance in
                the mass spectrum of the standard must be present in the
                mass spectrum of the sample component and should agree
                within absolute 10X.  For example, if an ion has a relative
                abundance of 30< in the standard spectrum, its abundance in
                the sample spectrum should be in the range of 20 to 40%.

         11.2.3 Identification is hampered when sample components are not
                resolved chromatographically and produce mass spectra
                containing ions contributed by more than one analyte.
                Because purgeable organic compounds are relatively small
                molecules and produce comparatively simple mass spectra,
                this is not a significant problem for most method analytes.
                When GC peaks obviously represent more than one sample
                component (I.e., broadened peak with shoulder(s) or valley
                between two or more maxima), appropriate analyte spectra and
                background spectra can be selected by.examining EICPs of
                characteristic ions for tentatively identified components.
                When analytes coelute (i.e., only one GC peak is apparent),
                the identification criteria described in Section 11.2.2 can
                be met but each analyte spectrum will contain extraneous
                Ions contributed by the coeluting compound.

         11.2.4 Structural isomers that produce very similar mass spectra
                can be explicitly Identified only if they have sufficiently
                different GC retention times.  Acceptable resolution 1s
                achieved if the height of the valley between two isomer
                peaks 1s less than 25X of the sum of the two peak heights.
                Otherwise, structural Isomers are identified as isomeric
                pairs.

12. CALCULATIONS

    12.1 Complete chromatographlc resolution 1s not necessary for accurate
         and precise measurements of analyte concentrations, if unique Ions
         with adequate Intensities are selected for EICPs.  For example,
         although two listed analytes, 1,1,2,2-tetrachloroethane and
         tetrachloroethene, were not resolved with the GC conditions used
         and produced mass spectra containing common Ions, concentrations
         (Table 5i were calculated by measuring appropriate characteristic
         Ions.

         12.1.1 With either the Internal or external standard technique,
                calculate analyte concentrations with the equation:


                      cx  •  V  Q!          •
                       *     AS .  RF  .  V

-------
                where Cx •   analyte concentration  in mlcrograms per  liter;
                      Ax »   integrated ion abundance of a significant
                             characteristic ion of  the sample analyte;
                      As a   integrated 'ion abundance of a significant
                             characteristic ion of  the standard (either
                             internal or external), in units consistent with
                             those used for the analyte ion abundance;
                      RF   - response factor (With  an external standard,
                             RF * 1, because the standard is the same
                             compound as the measured analyte.);
                      Qs *   quantity of internal standard added or quantity
                             of external standard that produced As, in
                             mlcrograms; and
                      V    * purged sample volume in liters.

         12.1.2 With the external standard technique, As 1s a point on the
                concentration calibration curve and Is the same number as
                AX* QS 1s the Quantity that produces As and also is
                obtained from the concentration calibration curve.

         12.1.3 For each analyte, select a significant characteristic ion.
                When feasible, use the most Intense ion in the mass
                spectrum; when a less Intense ion 1s more characteristic and
                sufficiently Intense to provide necessary sensitivity, use
                that ion to avoid possible Interferences.

13. PRECISION AND ACCURACY

    13.1 To obtain method efficiency data (Table 2) and to Indicate
         anticipated single laboratory accuracy and precision data (Table 5)
         for each listed analyte, seven 25-mL allquots of each of two
         solutions of reagent water containing known amounts of analytes
         were analyzed with purge and trap procedures and a packed column.
         One solution contained 16 ug of analyte per liter of solution; the
         other contained 1.6 ug/L.  Two direct injections of appropriate
         volumes of secondary dilution standard were Interspersed among
         purged allquots.  To obtain the data in Table 5, one aliquot of
         each of the two laboratory control standards was randomly selected
         to be a solution with known true values of analytes.  This aliquot
         was treated as an external standard, and the other six allquots of
         each of the two solutions were treated as  samples.

         13.1.1 Except for two listed analytes, mean method efficiency
                varied among analytes from 25.OX to 118.7X.  Those two
                aaalytes, l,2-dibromo-3-chloropropane and b1s(2-ch1oro-
                Isopropyl) ether, are very Inefficiently purged and were not
                detected in allquots containing 1.6 ug/L; mean method
                efficiencies for these two analytes when purged from  16 ug/L
                allquots were 9.4% and 4.3X, respectively (Table 2).
                Although for some applications these low efficiencies may
                result 1n unacceptably high detection limits for those

-------
       analytes, they can be measured with acceptable accuracy and
       precision when present at a concentration of 16 ug/L (Table
       5).

13.1.2 With these data, MOLs were calculated using the formula:


         ML ' ^n-l.l-  - 0.99)

where:
t(n-l, i_  « o.99) " Student's t value for the 9956 confidence
                       level with n-1 degrees of freedom, where
                       n - number of replicates, and
                 s * standard deviation of replicate analyses.

-------
REFERENCES

1.  Glaser, J. A., D. L. Foerst, G. 0. McKee, S. A. Quave, and W. L. Budde,
    "Trace Analyses for Wastewaters," Environ. Sc1_. Techno!. 15, 1426, 1981

2.  "Carcinogens - Working With Carcinogens," Department of Health,
    Education, and Welfare, Public Health Service, Center for Disease
    Control, National Institute for Occupational Safety and Health ,
    Publication No. 77-206, Aug. 1977.

3.  "OSHA Safety and Health Standards, General Industry," (29CFR1910),
    Occupational Safety and Health Administration, OSHA 2206, (Revised,
    January 1976).

4.  "Safety 1n Academic Chemistry Laboratories," American Chemical Society
    Publication, Committee on Chemical Safety, 3rd Edition, 1979.

5.  "The Determination of Halogenated Chemicals 1n Water by the Purge and
    Trap Method," Method 502.1,  EPA 600/4-81-059,  U.S. Environmental
    Protection Agency, Office of Research and Development, Environmental
    Monitoring and Support Laboratory, Cincinnati, OH, April 1981.

6.  "The Analysis of Aromatic Chemicals 1n Water by the Purge and Trap
    Method,"  Method 503.1, EPA 600/4-81-057,   U.S. Environmental
    Protection Agency* Office of Research and Development, Environmental
    Monitoring and Support Laboratory, Cincinnati, OH, May 1980.

-------
Table 2.  Single Laboratory Method Efficiency Data4 for  Purgeable
              Organic Compounds Measured with GC/MS
Compound
chloromethane

vinyl chloride

methyl ene chloride

1,1-d1 chl oroethene

bromochl oromethane

tr ans - 1 , 2 -d 1 ch 1 oroethene

chloroform

l,2-d1chloroethane

1,1,1 -tr1 chl oroethane

carbon tetrachloride

bromodl chloromethane

1 , 1 ,2-tr1 ch loroethane

tr1 chl oroethene

benzene

ch 1 orod 1 bromomethane

c1s-l,3-d1ch1oropropene

fluorobenzene

Measured
Ion
50

62

84

96

128

96

83

62

97

117

83

97

130

78

129

75

96

Rel.
Ret.
T1meb
0.10

0.14

0.22

0.28

0.30

0.31

0.44

0.50

0.59

0.62

0.66

0.79

0.84

0.88

0.88

0.90

1.00

True
Cone.
uq/L
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16 '
1.6
16
1.6
16
Mean Method
Efficiency
X

108.5

53.8
118.7
81.9
93.8
88.1
68.8
65.0
50. OC
98. 8C
93.8
92.5
62.5
55.6
93.8
90.6
106.2
110.0
81.3
80.6
18.8C
50.0C
100.0
106.9
106.2
96.3
58.8
53.8
25. OC
70. OC
100.0
95.6
Rel.
Std.
Oev., %

15.2

13.6
4.9
7.6
5.1
7.9
5.8
6.2
30. 5C
12.6C
3.8
2.1
12.6
7.1
3.8
2.9
2.2
1.9
5.9
5.4
33. 5C
15. 9C
3.0
1.2
2.9
2.2
6.0
11.3
31. 6C
11. 8C
7.4
11.1

-------
                               Table 2.   (continued)
Compound
bromoform

1,1,2,2-tetrachloroethane

tetrachloroethene

toluene

chlorobenzene

1 ,2-d1bromo-3-chloropropaned

4-bromof 1 uorobenzene

styrene

p-xylene

b1s(2-chloro1sopropyl) ether*1

l,3-d1chlorobenzene

1 ,2-d1 chlorobenzene

Measured
Ion
173

83

164

92

112

157

174

104

106

45

146

146

Rel.
Ret.
T1meb
1.10

1.29

1.31

1.42

1.52

1.70

1.82

1.93

1.97

2.08

2.19

2.20

True
Cone.
uq/L
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16

16
1.6
16
1.6
16
Mean Method
Efficiency
%
39.4
35.0
28.1
25.0
106.2
107.5
100.0
102.5
93.8
96.3
— d
9.4C
93.8
93.8
68. 8C
86. 9C
100.0
103.1

4.3
75. QC
82. 5C
81.3
79.4
Rel.
Std.
Dev., %
7.0
6.3
5.9
6.3
0.0
4.6
3.2
1.5
3.5
2.2
-d
10.3C
5.4
4.0
29. 2«
13. 3C
3.0
1.5

14.3
28. 3«
10.3C
5.9
4.2
a Except as noted, data were produced by purging seven  allquots of  reagent
  water spiked with known amounts of listed compounds;  calculations Involved  two
  direct Injections.
b GC column: 1.8 m x 2 mm ID glass packed with  IX  SP-1000 on 60/80  mesh
  carbopack B.  Program: 45°C for 4 m1n; 8°C/m1n to 230°C.
  Retention time relative to fluorobenzene, which  has retention time of  11.1  m1n
  under described GC conditions.
c Produced by analysis of six allquots rather than seven.
d Compound 1s very Inefficiently purged from water and  was  not detected  1n
  allquots of 1.6 ug/l solution.

-------
  Table  1.   Ion  Abundance  Criteria for £-Bromof1uorobenzene


 Mass	Ion Abundance Criteria	


 50                           15 to 40X of mass 95
 75                           30 to 60% of mass 95
 95                           Base Peak, 100X Relative Abundance
 96                           5 to 9% of mass 95
173                           < 2X of mass 174
174                           > 50% of mass 95
175                           5 to 9X of mass 174
176                           > 95X but < 101X of mass 174
177                           5 to 9X of mass 176

-------
                              Table 3 (continued)
                                       Demonstrated Acceptable
     Analyte                                Storage Time,
                                                 Days
2-chloroethylethyl ether                         27
2-chloroethylvinyl ether                         27
bis-2-chloroethyl ether                           9°
b1s-2-chloro1sopropyl ether                      27


a These data were obtained by multiple analyses  of raw  river  water  and
  carbon-filtered chlorinated tap water to which known  amounts  (0.20 to 0.50
  ug/L) of listed analytes had been added.  Some samples were stored and
  analyzed periodically over a 21-day period; others, over  a  27-day period.
  Data from "The Determination, of Halogenated Chemicals 1n  Water  by the
  Purge and Trap Method," Method 502.1, EPA 600/4-81-059, U.S.  Environmental
  Protection Agency, Office of Research and Development, Environmental
  Monitoring and Support Laboratory, Cincinnati, OH, April  1981.

D Because of observed changes during storage, this number 1s  the  maximum
  recommended storage time.

-------
          Table 3.  Acceptable Storage Times3 for River and Drinking
                   Water Samples Containing Halogenated Aliphatic Analytes
                                       Demonstrated Acceptable
     Analyte                                Storage Time,
                                                Days


chloromethane                                    21
dlchloromethane                                  27
carbon tetrachlorlde                             27
bromomethane                                      2°
dlbromomethane                                   21
bromoform                                        27
bromochloromethane                               21
bromodlchloromethane                             27
chlorodlbromomethane                             27
d1chlorod1fluoromethane                          27
fluorotrlchloromethane                           27

chloroethane                                     21
I,l-d1chloroethane                               27
1,2-d1chloroethane                               27
1,1,1-tri chloroethane                            21
1,1,2-tr1chloroethane                            27
1,1,1,2-tetrachloroethane                        21
pentachloroethane                                27
l,2-d1bromoethane                                21

chloroethylene (vinyl chloride)                   6b
I,l-d1chloroethylene                             27
c1s + trans-l,2-d1chloroethylene                 27
c1s-l,2-dichloroethylene                         21
I,l,2-tr1chloroethylene                          27
1,1,2,2-tetrachloroethylene                      27

l,2-d1chloropropane                              21
l,3-d1chloropropane                              21
1,2,3-tHchloropropane                           21

3-chloroprop-l-ene (allyl chloride)               2b
Ij-d1chloroprop-l-ene                           27
2,3-d1chloroprop-l-ene                            6"
trans-l,3-d1chloroprop-l-ene                      1?
c1s-l,3-d1chloroprop-l-ene                        1D

1-chlorohexane                                    lb
chlorocyclohexane                                27
1-chlorocyclohex-l-ene                           27

-------
                             Table 4.  Storage Time Oataa for River and Drinking
                                  Water Samples  Containing  Aromatic Analytes
Analyte
  Percent Recovery From
  River Water on Day 1
Preserved*1      Unpreserved
Acceptable0 Storage Time
        In Days
  Drinking     Preserved
   Water      River Water
    Observed
    Average Decay
   Rate (X/Day) In
Preserved River Water
benzene
toluene
ethylbenzene
o-xylene '
m-xylene
p-xylene
ethenylbenzene (styrene)
n-propylbenzene
Isopropylbenzene
n -butyl benzene
sec-butylbenzene
ter -butyl benzene
1 -methyl -4- Isopropylbenzene
1 , 2, 4 -trl methyl benzene
1 ,3,5-trlmethylbenzene
chlorobenzene
bromobenzene
o-dl chlorobenzene
m-d1 chl orobenzene
p-dl chlorobenzene
1 . 2, 3-trl chlorobenzene
1,2,4-trlchlorobenzene
o-chloro toluene
p-chlorotoluene
«- 1 r 1 f 1 uoroto 1 uene
100
98
95
95
93
88
83
88
93
88
88
90
88
85
88
90
90
96
96
94
86
88
—
90
84
12
8
10
12
17
12
0
15
0
3
10
43
34
16
5
23
17
85
96
90
97
88
_-
45
74
15
15
15
15
15
15
Qd
15
15
15«
15
15
6f
15*
28
28
28
28
28
28
28
28
21
13
28
14
14
—
—
—
—
—
—
—
—
—
—
—
—
—
26
26
26
26
26
26
26
—
26
26


1.4
1.6
1.6
2.3
4.2
2.2
1.5
3.2
2.2
10
2.3
3.0
2.1










a Data from "The Determination of Aromatic  Chemicals  In Water by the Purge and Trap Method," Method 503.1,
  EPA 600/4-81-057, U.S. Environmental Protection Agency,  Office of Research and Development, Environmental
  Monitoring and Support Laboratory, Cincinnati,  OH,  May 1980, and from "The Determination of Halogenated
  Chemicals In Water by the Purge and Trap  Method,* Method 502.1, EPA 600/4-81-059, U.S.  Environmental
  Protection Agency, Office of Research  and Development, Environmental Monitoring and Support Laboratory,
  Cincinnati, OH, April 1981.
b Preservation was accomplished by  adjustment  of  sample pH to 2.
c Data were obtained by multiple analyses of carbon-filtered tap water and preserved river water to which
  known amounts (0.40 or 0.50 ug/L) of listed  analytes had been added.   Mean recovery of analyte was >_ 80*
d Styrene, which reacts with free chlorine was  not  detected In chlorinated drinking water.
f Mean recovery of 78X.
f Mean recovery of 75X.

-------
Table 5.  Anticipated Accuracy and  Precision  Data with Method
            (Single Laboratory and Single Operator)
Compound
chloromethane i

vinyl chloride

methylene chloride

1 , 1 -d 1 ch 1 or oe thene

bromochl or ome thane

trans-1 ,2-dlchloroethene

chloroform

1,2-dlchloroethane

1 , 1 . 1- tr1 chl oroethane

carbon tetrachlorlde

bromodi chl or ome thane

1 ,1 ,2-tr1chloroethane

trlchloroethene

benzene

ch 1 orodi bromome thane

ds-l,3-d1chloropropene

fluoro benzene
Measured
Ion
50

62

84

96

128

96

83

62

97

117

83

97

130

78

129

75

96
Rel.
Ret.
T1meb
0.10

0.14

0.22

0.28

0.30

0.31

0.44

0.50

0.59

0.62

0.66

0.79

0.84

0.88

0.88

0.90

1.00
True
Cone.
ug/L
oe
16
oe
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16

Mean
Observed
Cone., yg/L

16.5

18.2
1.6
15.8
1.6
16.3
1.6
15.6
1.3
15.4
1.6
16.0
1.6
16.0
1.6
16.2
1.6
16.0
1.6
15.8
1.3
15.9
1.6
16.0
1.6
16.1
1.6
15.0
1.3
15.0

Std.
Dev.
ug/L

2.7

2.5
0.08
0.3
0.08
1.8
0.08
1.4
0.5
2.2
0.06
0.32
0.1
1.7
0.04
0.8
0.04
0.4
0.09
0.5
0.3
2.5
0.05
0.2
0.06
0.5
0.1
2.0
0.5
1.8

Rel.
Std.
Dev.,*

16.6

13.9
4.8
2.1
4.9
10.8
5.2
9.2
37.9
14.5
3.8
2.0
6.7
10.5
2.4
4.9
2.3
2.4
5.5
3.2
23.1
15.9
3.3
1.1
4.0
3.1
6.5
13.6
36.9
11.7

Mean Method
Accuracy
%

103.1

113.8
100
98.8
100
101.9
100
97.5
81.3
96.3
100
100
100
100
100
101.3
100
100
100
98.8
81.3
99.4
100
100
100
100.6
100
93.8
81.3
93.8

Method
Detection
L1m1tc, ug/L

9.2

8.5
0.25
~
0.27
—
0.28
—
1.7
—
0.20
—
0.35
—
0.13
—
0.13
—
0.29
—
1.0
—
0.18
—
0.21
—
0.34
—
1.8
—


-------
                                                 Table 5.   (continued)
Compound
bromoform

1 ,1 ,2,2-tetrachloroethane

tetrachloroethene »

toluene

chlorobenzene

1 ,2-d1bromo-3-chloropropaned

4-bromofluorobenzene

styrene

p-xylene

b1s(2-chloro1sopropy1) etherd

1 ,3-d1chlorobenzene

1 , 2 -di chlorobenzene

Measured
Ion
173

83

164

92

112

157

174

104

106

45

146

146

Rel.
Ret.
T1meb
1.10

1.29

1.31

1.42

1.52

1.70

1.82

1.93

1.97

2.08

2.19

2.20

True
Cone.
ug/L
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
Mean
Observed
Cone.. ug/L
1.5
15.5
1.6
15.5
1.6
15.8
1.6
15.9
1.6
15.7
_.d
15.7
1.5
15.6
1.2
15.8
1.6
16.1
-Id
15.1
1.2
15.2
1.5
15.7
Std.
Dev.
U9/L
0.1
2.0
0.08
1.8
0.02
0.4
0.02
0.3
0.03
0.6
-_d
3.0
0.09
1.1
0.3
1.6
0.05
0.2
— d
2.3
0.3
1.5
0.09
1.3
Rel. Mean Method
Std . Accuracy
Dev.,* %
6.5
12.7
5.3
11.5
1.3
2.4
1.4
1.6
1.6
3.6
..d
19.1
5.6
7.3
29.0
10.4
3.4
1.0
..d
15.1
28.0
9.8
6.2
8.3
93.8
96.9
100
96.9
100
98.8
100
99.4
100
98.1
—d
98.1
93.8
97.5
75.0
98.8
100
100.6
__d
94.4
75.0
95.0
93.8
98.1
Method
Detection
L1m1tc,uq/L
0.34
—
0.28
—
0.07

0.08
—
0.09
__
_.d
j
10. ld
0.29
—
1.3
—
0.18
— _
__d
~A
8.6d
1.3
-
0.30
-
a Produced by analysis of  seven  allquots  of reagent water spiked with known amounts of listed compounds;
  calculations based on external standard technique.  Two allquots were treated as standards; five
  allquots were treated as  samples.
b GC column:  1.8 m x 2 mm ID  glass  packed with IX SP-1000 on 60/80 mesh Carbopack B.  Retention times
  relative to fluorobenzene.
c Minimum concentration that can be  measured with 99* confidence that reported value Is greater than
  zero.*
d Compound Is Inefficiently purged and  was not  detected 1n 1.6 ug/L solution; MDL was calculated from
  analysis of 16 ug/L solution.
6 Compound not analyzed at  concentration  of 1.6 ug/L.

-------