r  Method 505!  Analysis of Organohallde
  Pesticides and Aroclors 1n Drinking Mater
  by Mlcroextractlon and Gas Chroaatography
                September 1986

 Suppleaent to "Methods for the Determination
  of Organic Compounds  1n  Finished  Drinking
         Water  and Raw  Source Water*
     Physical and Chemical  Methods Branch
Environmental Monitoring and Support Laboratory
     U. S. Environmental Protection Agency
             Cincinnati, Ohio 45268

-------
       KTHOO  505.   ANALYSIS OF ORGANOHAUOE PESTICIDES AND AROCLORS  IN
           DRINKING WATER BY HICROEXTRACTION AND GAS CHROHATOGRAPHY


1.  SCOPE AND APPLICATION

    1.1  This method (1,2,3) 1s Applicable to the determination of the
         following analytes 1n finished drinking water, drinking water
         during Intermediate stages of treatment, and the raw source water:

             Analyte                                CAS No.

         Alachlor                                15972-60-8
         Aldrln                                    309-00-2
         Chlordane                                  57-74-9
         D1e1dr1n                                   60-57-1
         EndHn                                     72-20-8
         Heptachlor                                 76-44-8  '
         Heptachlor Epoxlde                       1024-57-3
         Hexachlorobenzene                         118-74-1
         Llndane                                    58-89-9
         Methoxychlor                               72-43-5
         Toxaphene                                8001-35-2
         Aroclor 1016                            12674-11-2
         Aroclor 1221                            11104-28-2
         Aroclor 1232                            11141-16-5
         Aroclor 1242                            53469-21-9
         Aroclor 1248                            12672-29-6
         Aroclor 1254                            11097-69-1
         Aroclor 1260                            11096-82-5

    1.2  PCBs are determined as Aroclors by this method.

    1.3  For compounds other than the above mentioned analytes or for other
         sample sources, the analyst must demonstrate the applicability of
         the method by collecting precision and accuracy data on actual
         samples (4) and provide qualitative confirmation of results by Gas
         Chromatography/Mass Spectrometry (GC/MS) (5), or by GC analysis
         using dissimilar columns.

    1.4  The experimentally determined method detection limits (MDL) (6) for
         the above organohalldes and Aroclors are shown 1n Table 1.  Actual
         detection limits are  highly dependent upon the characteristics of
         the gas chromatographlc system used (e.g. coluon type, age, and
         proper conditioning;  detector condition; and Injector mode and
         condition).

2.  SUMMARY OF METHOD

    2.1  Thirty-five mL of  sample are extracted wltfi 2 aL of hexane.  Two uL
         of the extract are then Injected Into a gas chromatograph equipped
         with a linearized  electron capture detector for separation and

-------
         analysis.  Aqueous calibration standards are extracted and  analyzed
         1n an Identical manner 1n order to compensate for possible
         extraction losses.

    2.2  The extraction and analysis time 1s 30 to 50 minutes per sample
         depending upon the analytes and the analytical conditions chosen.
         (See Tables 1-2 and Figures MO.)

    2.3  Confirmatory evidence can be obtained using a dissimilar column.
         When component concentrations 1n water samples are sufficiently
         high, SC/MS may be employed.

3.  INTERFERENCES

    3.1  Impurities contained 1n the extracting solvent usually account for
         the majority of the analytical problems.  Solvent blanks should be
         analyzed on each new bottle of solvent before use.  Indirect dally
         checks on the extracting solvent are obtained by monitoring the
         laboratory reagent blanks (9.1.1).  Whenever an Interference 1s
         noted 1n the sample blank, the analyst should analyze another
         solvent blank.  Low level Interferences generally can be removed by
         distillation or column chromatography (3); however, 1t 1s generally
         more economical to obtain a new source solvent.  An
         Interference-free solvent 1s a solvent containing non-detectable
         peaks at the retention times of the analytes of Interest.

    3.2  Caution must be taken 1n the determination of endrln since  1t has
         been reported that the splitless Injector may cause endrln
         degradation (7).  The analyst should be alerted to this possible
         Interference resulting 1n an erratic response for endrln.

    3.3  Variable amounts of pesticides and Aroclors from aqueous solutions
         adhere to glass surfaces.  It 1s recommended that sample transfers
         and glass surface contacts be minimized.

    3.4  Aldrin and methoxychlor are rapidly oxidized by chlorine.
         Dechlor1nat1on with sodium thlosulfate at time of collection will
         retard further oxidation of these compounds.

4.  SAFETY

    4.1  The toxldty and carc1nbgen1c1ty of chemicals used In this  method
         have not been precisely defined; each chemical should be treated as
         a potential health hazard, and exposure to these chemicals  should
         be minimized.  Each laboratory is responsible for maintaining
         awareness of OSHA regulations regarding safe handling of chemicals
         used in this method.  Additional references to laboratory safety
         are available (9-10) for the information of. the analyst.

    4.2  The following organohalides have been tentatively classified as
         known or suspected human or mammalian carcinogens:  aldrin.

-------
         Aroclors, chlordane, dleldrln, heptachlor, hexachlorobenzene, and
         toxaphene.  Pure standard materials and stock standard solutions of
         these compounds should be handled  1n a hood or glovebox.  A NIOSH/.
         MESA approved toxic gas respirator should be worn when the analyst
         handles high concentrations of these toxic compounds.

5.  APPARATUS AND EQUIPMENT

    5.1  SAMPLE CONTAINERS - 40-ot screw cap vials (Pierce #13075 or
         equivalent) each equipped with a PTFE-faced slllcone septum (Pierce
         #12722 or equivalent).  Prior to use, wash vials and septa with
         detergent and rinse with tap and distilled water.  Allow the vials
         and^septa to air dry at room temperature, place the vials 1n a
         400*C oven for one hour, then remove and allow to cool 1n an area
         known to be free of organlcs.

    5.2  VIALS - auto sampler, screw cap with septa, 1.8 «L, Varlan
         #96-000099-00 or equivalent or any other autosaapler vials not
         requiring more than 1.8 oL sample  volumes.

    5.3  AUTO SAMPLER - Hewlett-Packard 7671A, or equivalent.

    5.4  MICRO SYRINGES - 10 and 100 VL.

    5.5  MICRO SYRINGE - 25 VL with a 2-1nch by 0.006-Inch needle - Hamilton
         702N or equivalent.

    5.6  PIPETTES - 2.0 and 5.0 mL transfer.

    5.7  VOLUMETRIC FLASKS - 10 and 100 mL, glass stoppered.

    5.8  STANDARD SOLUTION STORAGE CONTAINERS - 15-nL bottles with
         PTFE-Uned screw caps.

    5.9  GAS CHROMATOGRAPHY SYSTEM

        5.9.1  The GC must be capable of  temperature programming and should
               be equipped with a linearized electron capture detector,
               capillary column, and splUless  Injector (0.5 m1n. spHtless
               mode).  Alternately, an on-co1umn  Injector may be employed.

        5.9.2  Three gas chromatographlc  columns  are recommended.  Column  A
               should be used as the primary analytical column unless
               routinely occurring  analytes are not adequately resolved.
               Columns B and C  are  recommended for use as confirmatory
               columns when GC/MS confirmation  1s not available.

               5.9.2.1  Column  A -  0.32mm 10 x  30 M .long fused silica
                        capillary with chemically bonded methyl  polysiloxane
                        phase  (DB-1,  l.Owm film,  or-equivalent).  The  linear
                        velocity of the  helium carrier gas  is established  at
                        25 on/sec.  The  column  temperature  is programmed

-------
                         from  180'C to 260*C  at 4'C/m1n  and held at 260*C
                         until all expected compounds have e luted.  Injector
                     -    temperature:  200'C.  Detector  temperature:  290*C.*
                         (See Table 1 for retention data.)

               5.9.2.2   Column B  (alternate  confirmation column) •- 0.32mm 10
                         x 30 M long fused silica capillary with a 1:1 nixed
                         phase of  dimethyl s111cone and  polyethylene glycol
                         (Durawax-OX3, 0.25vm film, or equivalent).  The
                         column temperature 1s programmed from 100*C to 210*C
                         at 8'CMIn, and held at 210*C until all expected
                         compounds have e luted.  Then the post temperature 1s
                         programed to 240*C  at 8*C/m1n  for 5 m1n.  (See
                         Table 2 for retention data).

               5.9.2.3   Column C  (alternate  confirmation column) - 0.32mm 10
                         x 25 M long fused silica capillary with chemically
                         bonded 50:50 Methyl-Phenyl $111 cone (OV-17, 1.5U
                         film thickness, or equivalent). The linear velocity
                         of the helium carrier gas 1s established at 39
                         cm/sec.   The column  temperature 1s programmed from
                         100'C to  260*C at 4*C/m1n and held at 260*C until
                         all expected compounds have e luted.

6.  REAGENTS AND CONSUMABLE MATERIALS

    6.1  REAGENTS

         6.1.1  Hexane  extraction solvent -  UV Grade, Burdlck and Jackson
                     or equivalent.
         6.1.2  Methyl  alcohol - ACS Reagent Grade,  demonstrated  to be free
                of  analytes.

         6.1.3  Sodium  chloride, NaCl - ACS Reagent  Grade  - For pretreatment
                before  use, pulverize a batch of  NaCl  and  place 1n a muffle
                furnace at room temperature.  Increase the temperature to
                400* C and hold for  30 m1n.  Place 1n a bottle  and cap.

         6.1.4  Sodium  thlosulfate, NagS903 — ACS  Reagent Grade— For
                preparation of solution (0.04 g/mL), mix 1 g of
                Na2^2^3 w
-------
    6.3  STANDARD STOCK SOLUTIONS - These solutions wy be obtained as
         certified solutions or prepared from pure standard materials using
         the following procedures:

         6.3.1  Prepare stock  standard solutions (5000 ug/mL) by accurately
                weighing about 0.0500 g of pure material.  Dissolve the
                material In nethanol and dilute to volume 1n a 10-mL
                volumetric flask.  Larger volumes can be used at the
                convenience of the analyst.  When compound purity 1s assayed
                to be 961 or greater, the weight can be used without
                correction to  calculate the concentration of the stock
                standard.  Commercially prepared stock standards can be used
                at any concentration 1f they are certified by the
                manufacturer or by an Independent source.

         6.3.2  Transfer the stock standard solutions Into Teflon-sealed
                screw-cap bottles.  Store at 4*C and protect from light.
                Stock standard solutions should be checked frequently for
                signs of degradation or evaporation, especially just prior
                to preparing calibration standards from them.

         6.3.3  Stock standard solutions must be replaced after six months,
                or sooner 1f comparison with check standards Indicates a
                problem.

    6.4  SECONDARY DILUTION STANDARDS — Use standard stock solutions to
         prepare secondary dilution standard solutions that contain the
         analytes 1n methane!. The secondary dilution standards should be
         prepared at concentrations that can be easily diluted to prepare
         aqueous calibration standards (Sect. 8.1.1) that will bracket the
         working concentration range.  Store the secondary dilution standard
         solutions with minimal headspace and check frequently for signs of
         deterioration or evaporation, especially just before preparing
         calibration standards.  The storage time described for stock
         standard solutions 1n Sect. 6.4.3 also applies to secondary
         dilution standard solutions.

7.  SAMPLE COLLECTION. PRESERVATION. AND STORA6E

    7.1  SAMPLE COLLECTION

         7.1.1  Collect all samples 1n duplicate 40-mL bottles Into which 3
                rag of sodium thlosulfate crystals have been added to the
                empty bottles  just prior to shipping to the sampling site.
                Alternately, 75 wL of freshly prepared sodium thlosulfate
                solution (0.04 g/nt) may be added to empty 40-mL bottles
                just prior to  sample collection.  In collecting field
                samples, 1t is recommended to add sodium thlosulfate
                solution at the sampling site.

-------
         7.1.2  When sampling from a water tap, open  the tap  and  allow  the
                system to flush until the water temperature has stabilized
                (usually about 10 min).  Adjust the flow to about 500 mL/m1n
                and collect duplicate samples from the  flowing stream.

         7.1.3  When sampling from a well, fill a w1de-aouth  bottle or
                beaker with sample, and carefully fill  duplicate  40-ot
                sample bottles.

    7.2  SAMPLE PRESERVATION

         7.2.1  The samples must be chilled to 4*C at the time of collection
                and maintained at that temperature until the  analyst 1s
                prepared for the extraction process.  Field samples that
                will not be received at the laboratory  on the day of
                collection must be packaged for shipment with sufficient  1ce
                to Insure that they will be maintained  at 4*C until arrival
                at the laboratory.

    7.3  SAMPLE STORAGE

         7.3.1  Store samples and extracts at 4*C until extraction and
                analysis.

         7.3.2  Extract all samples as soon as possible after collection.
                Results of holding time studies suggest that  all  analytes
                (aldrln, dleldrln, endrln, heptachlor epoxlde, hexachloro- .
                benzene, Undane, and the aroclors) with the  possible
                exception of heptachlor may be extracted within 14 days
                after collection.  In general, heptachlor showed
                Inconsistent results.  If heptachlor  1s to be determined,
                samples must be extracted within 7 days of collection.
                Samples from which analytes have not  been extracted within
                these prescribed periods of time must be discarded and
                replaced.

8.  CALIBRATION AND STANDARDIZATION

    8.1  CALIBRATION

         8.1.1  At least three calibration standards  are needed.   One  should
                contain analytes at a concentration  near but  greater than
                the method detection limit for each  compound; the other two
                should be at concentrations that bracket the  range expected
                1n samples.  For example, If  the MOL  1s 0.01  ug/L, and  a
                sample expected to contain approximately 0.10 ug/L 1s  to be
                analyzed, aqueous standards should  be prepared at
                concentrations of 0.02 ug/L,  0.10 ug/L, and 0.20  ug/L.
                                                   ;^
         8.-1.2  To prepare a calibration standard,  add  an appropriate  volume
                of a secondary dilution standard to  a 35-
-------
                    water 1n a 40-mL bottle.  Do not add less than 20 UL
            of an alcoholic standard to the reagent water.   Use a 25-uL
            micro syringe and rapidly Inject the alcoholic  standard Into
            the middle point of the water volume.  Remove the needle as
            quickly as possible after Injection.  Mix by Inverting and
            shaking the capped bottle several times.  Aqueous standards
            must be prepared fresh dally.

     8.1.3  Starting with the standard of lowest concentration, prepare,
            extract, and analyze each calibration standard  beginning
            with Sect. 10.1.4 and tabulate peak height or area response
            versus the concentration 1n the standard.  The results are
            to be used to prepare a calibration curve for each compound
            by plotting the peak height or area response versus the
            concentration.  Alternatively, 1f the ratio of  response to
            concentration (calibration factor) 1s a constant over the
            working range (<10X relative standard deviation), linearity
            to the origin can be assumed and the average ratio or
            calibration factor can be used 1n place of a calibration
            curve.

     8.1.4  The working calibration curve or calibration factor must be
            verified on each working day by the measurement of one or
            more calibration standards.  If the response for an analyte
            varies from the predicted response by more than *15X, the
            test must be repeated using a fresh calibration standard.
            If the results still do not agree, generate a new calibra-
            tion curve or use a single point calibration standard as
            described 1n Sect. 8.1.5.

     8.1.5  Single point calibration 1s an acceptable alternative to a
            calibration curve.  Prepare single point standards from the
            secondary dilution standard solutions.  The single point
            calibration standard should be prepared at a concentration
            that produces a response close (*20X) to that of the
            unknowns.  Do not use less than 20 uL of the secondary
            dilution standard solution to produce a single point
            calibration standard 1n reagent water.

8.2  INSTRUMENT PERFORMANCE - Check the performance of the entire
analytical system dally using data gathered from analyses of reagent
blanks, standards, duplicate samples, and the laboratory control
standard (Sect. 9.2.2).

     8.2.1  Significant peak tailing 1n excess of that shown for the
            target compounds in the method chromatograms (Figures 1-10)
            must be corrected.  Tailing problems are generally traceable
            to active sites on the GC column, improper column
            Installation, or the detector operation.

     8.2.2  Check the precision between replicate analyses.  A properly
            operating system should perform with an average relative

-------
                standard deviation of less than  10*.  Poor precision 1s
                generally traceable to pneumatic  leaks, especially at the
                Injection port.   If the 6C system 1s  apparently performing -
                acceptably but with decreased sensitivity, It may be
                necessary to generate a new curve or  set of calibration
                factors to verify the decreased  responses.

         8.2.3  Observed relative area responses  of endrin must meet the
                following criteria:  endrin £ 501 of  total area.

9.  QUALITY CONTROL

    9.1 MONITORING FOR INTERFERENCES

         9.1.1  Laboratory Reagent Blanks - A laboratory reagent blank 1s an
                aliquot of reagent water analyzed as  1f 1t were a sample.
                Add NgSjOa to laboratory reagent blanks.  Analyze a
                laboratory reagent blank each day and as necessary to
                Identify sources  of contamination. The laboratory reagent
                blank should contain less than the HM. response of each
                analyte.

    9.2  ASSESSING ACCURACY

         9.2.1  Each quarter, 1t  1s essential that the laboratory analyze
                quality control check samples for each contaminant.  If any
                criteria established by USEPA are not met, corrective action
                needs to be taken and documented.

         9.2.2  After every 10 samples, and preferably 1n the middle of each
                day, analyze a laboratory control standard.  Calibration
                standards may not be used for accuracy assessments and the
                laboratory control standard may  not be used for calibration
                of the analytical system.

                9.2.2.1   Laboratory Control Standard Concentrate - If
                          Internally prepared laboratory control standards
                          are used to provide the routine assessment of
                          accuracy, they should  be prepared from a separate
                          set of  stock standards. Fran stock standards
                          prepared as described  1n Sect. 6.4, add a suffi-
                          cient volume of each stock  standard to methane1 1n
                          a 10-mL volumetric flask to yield a concentration
                          of 2.5  ug/mL and adjust to  volume.

                9.2.2.2   Laboratory Control Standard — Add 20 uL of the
                          control concentrate to a 35-*. aliquot of reagent
                          water poured Into a 40-raL'bottle containing 75 ul
                          of 0.04 g/mL sodium thlosulfate.

                9.2.2.3   Analyze the 35-mL aliquot as described In Sect.
                          10.  For each analyte  In the laboratory control

-------
                          standard,  calculate  the percent recovery  (Pj)
                                the equation:
                          where S-f  • the  analytical result from the
                                        laboratory control standard,  1n ug/L;
                                        and
                                 TI  •   the  known concentration of the spike,
                                        1n »g/L.

         9.2.3  It  1s essential  that the  laboratory analyze an unknown
                performance  evaluation  sample  (when available) once  per year
                for all regulated contaminants measured.  Results need to be
                within acceptance limits  established by USEPA for each
                analyte.

10. PROCEDURE

    10.1 SAMPLE PREPARATION

         10.1.1 Remove samples  from storage and allow  them to equilibrate to
                room temperature.

         10.1.2 Remove the container caps.   Withdraw and discard a 5-mL
                volume using a  transfer plpet. Replace the container caps
                and weigh the containers  with  contents to the nearest 0.1 g
                and record these weights  for subsequent sample volume
                determinations  (Sect. 10.3).

         10.1.3 Remove the container cap  of each sample, and add 6 g NaCl
                (Sect. 6.1.3) to the sample bottle.  Recap and dissolve the
                NaCl by Inverting and shaking  the bottles several times
                (approx. 20  sec).

    10.2 EXTRACTION ANO ANALYSIS

         10.2.1 Remove the cap, and using a transfer or automatic dispensing
                plpet, add 2.0  mL of hexane.  Recap and shake vigorously by
                hand for l.mln. Invert the bottle and allow the water and
                hexane phases to separate.

         10.2.2 Remove the cap  and  carefully transfer  approximately  0.5 mL
                of  hexane layer Into an autosampler vial using a disposable
                glass plpet.
                                                  /^
         10.2.3 Transfer the remaining  hexane  phase, being careful not to
                Include any  of  the  water  phase, Into^a second autosampler
                vial.  Reserve  this second  vial at 4*C for an Immediate
                reanalysls If necessary.

-------
         10.2.4 Transfer  the first sample vial  to  an  autosampler set up to
                Inject  1-2 ul portions Into the gas chroaatograph for
                analysis  (See Sect. 5.9 for SC  conditions).  Alternately,
                1-2 ul  portions of samples, blanks, and standards nay be
                manually  Injected, although an  autosampler 1s strongly
                recommended.

    10.3 DETERMINATION  OF SAMPLE VOLUME IN BOTTLES NOT CALIBRATED

         10.3.1 Discard the remaining sanple/hexane Mixture from the sample
                bottle.  Shake off the remaining few  drops using short,
                brisk wrist movements.

         10.3.2 Rewelgh the empty container with original cap and calculate
                the net weight of sample by difference to the nearest 0.1 g
                (Sect.  10.1.2 Blnus Section 10.3.2).  This net weight (1n
                grams)  1s equivalent to the volume of water (In mL)
                extracted (Sect. 11.3).

11. CALCULATIONS

    11.1 Identify the organohalldes In the sample  chromatogram by comparing
         the retention  time of the suspect peak to retention times generated
         by the calibration standards and the laboratory control standard.
         Identify the multlcomponent compounds  using  all peaks that are
         characteristic of the specific compound from chromatograms
         generated with Individual standards.   Select the most sensitive and
         reproducible peaks for calculation purposes.

    11.2 Use the calibration curve or calibration  factor (Sect. 8.1.3) to
         directly calculate the uncorrected concentration (C^) of each
         analyte 1n the sample (e.g., calibration  factor x response).

    11.3 Calculate the  sample volume (Vs) as equal to the net sample
         weight:

           Ys - gross weight (Sect. 10.1.2) - bottle  tare (Sect. 10.3.2).

    11.4 Calculate the  corrected sample concentration as:

           Concentration, »g/L • C^  X   y


    11.5 Report the results for the unknown samples 1n Bg/L.  Round off the
         results to the nearest 0.1 ug/L or two significant figures.

12. ACCURACY AND PRECISION

    12.1 Single laboratory (EMSl-C1nc1nnat1) accuracy and precision at
         several concentrations in reagent, ground,'and tap water matrices
         are presented  In Table 3 (11).  These  results were obtained from
         data generated with a DB-1 column.

-------
               Table Z.  Chromatographlc Conditions and Retention Data
                        of  Organohallde  Pesticides  and  Aroclors
Analyte
Alachlor
Aldrln
Butachlor
Chlorpyrlfos
Chlorpyrlf os-Methyl
01ch1oben1l
Olchlorvos
01eldr1n
Endrln
Heptachlor
Heptachlor Epoxlde
Hexachl orobenzene
Lindane
Hetolachlor
Methoxychlor
Propachlor
Aroclor 1016

Aroclor 1221

Aroclor 1232

Aroclor 1242

Aroclor 1248

Aroclor 1254

Aroclor 1260

Chlordane

Toxaphene

Column B
Retention T1mel. M1n
19.71
18.37
25.16
21.52
10.46,
11.20
8.91
45.08
33.30
17.53
24.61
13.35
18.35
21.42
14
15
20
10

15
20
15
20
16
22
20
35
26
42
16
26
27

.43
.11.
.64
.46,

.10,
.62,
.10.
.63,
.60,
.66,
.55,
.72,
.56,
.33,
.28,
.28
.68,

21

16

11

16
22
16
22
19
23
25
36
31
43
17

35

.02

.62.

.62,

.61,
.68,
.61,
.66.
.42,
.48,
.63,
.38,
.21,
.25,
.51.

.56,



19.45,

13.19

19.43,
25.61
19.44,
25.64
20.41,
24.20,
26.62,
44.80
33.72,
44.68
24.57,

40.67,



20



20

20

20
25
32

36

25

43

Column C
Retention T1mel. M1n


.43



.42

.42

.61
.62
.99

.29

.60

.29

21
21
26
24
22
8
1
27
29
19
24
15
18
23
36
14
10
22
8
10
16
22
16
22
20
25
25
30
10
12
19
25
28
32
.09
.39
.75
.36
.35
.09
.18,
.76
.17
.99
.63
.62
.67
.50
.44
.70
.27,
.63.
.36,
.73,
.23,
.61,
.24,
.59.
.69.
.09,
.84,
.42.
.92.
.98.
.03,
.67,
.80.
.16,
5.48

18.52,
23.84
8.54,
12.78
18.49,
23.83,
18.48,
23.80
22.59,
25.56,
28.73,
32.14
11.53,
14.55,
20.00,
25.96,
29.25,
34.40


20.75,

10.27,

20.71
25.63
20.66

23.79,
26.75
29.67

12.29,
15.48
22.84,
26.32
30.33

!More than one peak listed does not Implicate the total number of peaks
characteristic of the multi-component pesticide,  listed peaks Indicate only the
ones chosen for quantification.

-------
13. REFERENCES

1.  Glaze, W.W., L1n, C.C., Optimization of Liquid-liquid Extraction Methods
    for Analysis of Organlcs 1n Water, EPA-600/S4-83-052, January 1984.

2.  Henderson, J.E., Peyton, S.R. and Glaze, U.K. (1976). In "Identification
    and Analysis of Organic Pollutants 1n Water* (L.H.  Keith ed.),
    pp. 105-111.  Ann Arbor Sc1. Publ., Ann Arbor, Michigan.

3.  Richard, J.J., Junk, G.A., 'Liquid Extraction for Rapid Determination of
    Halomethanes 1n Water,' Journal AWWA. 69, 62, January 1977.

4.  'Handbook for Analytical Quality Control 1n Water and Wastewater
    Laboratories,' EPA-600/4-79-019, U. S. Environmental Protection Agency,
    Environmental Monitoring and Support Laboratory, Cincinnati, Ohio,
    45268, March 1979.

5.  Budde, W.L., Elchelberger, J.W., "Organic Analyses Using Gas
    Chromatography-Mass Spectrometry,' Ann Arbor Science, Ann Arbor,
    Michigan 1979.

6.  Glaser, J.A. et al.. Trace Analyses for Wastewaters,' Environmental
    Science and Technology. 15, 1426 (1981).

7.  Bellar, T.A., Stemmer, P., Llchtenberg, J.J., "Evaluation of Capillary
    Systems for the Analysis of Environmental Extracts," EPA-600/S4-84-004,
    March 1984.

8.  'Carcinogens-Working with Carcinogens,' Department of Health, Education,
    and Welfare, Public Health Service, Center for Disease Control, National
    Institute of Occupational Safety and Health, Publication No, 77-206,
    August, 1977.

9.  'OSHA Safety and Health Standards, General Industry," (29CFR1910),
    Occupational Safety and Health Administration, OSHA 2206, (Revised,
    January 1976).

10. "Safety In Academic Chemistry Laboratories," American Chenrlcal Society
    Publication, Committee on Chemical Safety, 3rd Edition, 1979.

11. W1nf1eld, T., et al. "Analysis of Organohallde Pesticides 1n Drinking
    Water by Mlcroextraction and Gas Chromatography."   In preparation.

-------
       Table 1,  Chromatographlc Conditions and Method Detection Limits
                 of Organohallde Pesticides and Aroclors
                                    Column A
   Analyte                   Retention Time*. M1n         MQL.  Mq/L

Hexachlorobenzene           11.9                               0.002
Llndane                     12.3                               0.003
Alachlor                    1S.1                               0.22S
Heptachlor                  15.9                               0.003
AldHn                      17.6                               0.007
Heptachlor Epoxlde          19.0                               0.004
D1e1dr1n                    22.1                               0.012
EndHn                      24.2                               0.063
Methoxychlor                30.0                               0.956
Aroclor 1016                13.6, 14.8, 15.2                   0.08
                            16.2. 17.7
Aroclor 1221                 7.7, 9.0, 15.9, 19.1,            15.0
                            24.7
Aroclor 1232                ll'.2, 14.7. 13.6. 15.2             0.48
                            17.7
Aroclor 1242                11.2. 13.6, 14.7, 15.2             0.31
                            17.7, 19.8
Aroclor 1248                14.8, 16.2, 17.1, 17.7             0.102
                            19.8, 22.0
Aroclor 1254                19.1. 21.9. 23.4. 24.9             0.102
                            26.7
Aroclor 1260                23.4, 24.9, 26.7, 28.2             0.189
                            29.9. 32.6
Chlordane                   15.1, 15.9, 20.1, 20.9             0.14
                            21.3
Toxaphene                   21.7, 22.5, 26.7, 27.2             1.0


iMore than one peak listed does not Implicate the total number of peaks
  characteristic of the multi-component pesticide.  Listed peaks Indicate
  only the ones chosen for quantification.

Column A conditions:  Column A - 0.32mm ID x 30 M long fused silica
capillary with methyl polyslloxane phase (OB-1, 0.25 ua film, or
equivalent).  The linear velocity of the helium carrier gas Is established
at 25 cm/sec.  The column temperature 1s programmed fro« 180*C to 260*C at
4"C/min and held at 260*C until all expected compounds have eluted.
Injector temperature:  200*C; detector temperture:  290*C.

-------
               Table 2.  Chromatographlc Conditions and Rententlon Data
                        of Organohallde Pesticides  and  Aroclor*
                                    Column B                       Column C
   Analyte	Retention T1mel. M1n	Retention Tlmgl. M1n
Alachlor
Aldrln
Butachlor
Chlorpyrlfos
Chlorpyrlf os-Methyl
D1chloben1l
Olchlorvos
D1eldr1n
Endrln
Heptachlor
Heptachlor Epoxlde
Hex ach 1 orobenzene
Llndane
Metolachlor
Methoxychlor
Propachlor
Aroclor 1016

Aroclor 1221

Aroclor 1232

Aroclor 1242

Aroclor 1248

Aroclor 1254

Aroclor 1260

Chlordane

Toxaphene

19.71
18.37
25.16
21.52
10.46, 21.02
11.20
8.91
45.08
33.30
17.53
24.61
13.35
18.35
21.42
14.43
15.11, 16.62, 19.45, 20.43
20.64
10.46. 11.62, 13.19

15.10. 16.61, 19.43, 20.42
20.62, 22.68, 25.61
15.10, 16.61, 19.44, 20.42
20.63, 22.66, 25.64
16.60, 19.42, 20.41, 20.61
22.66, 23.48, 24.20, 25.62
20.55, 25.63, 26.62, 32.99
35.72, 36.38, 44.80
26.56, 31.21, 33.72, 36.29
42.33, 43.25. 44.68
16.28, 17.51, 24.57, 25.60
26.28
27.68, 35.56, 40.67, 43.29

21.09
21.39
26.75
24.36
22.35
8.09
1.18, 5.48
27.76
29.17
19.99
24.63
15.62
18.67
23.50
36.44
14.70
10.27, 18.52, 20.75,
22.63, 23.84
8.36, 8.54, 10.27,
10.73, 12.78
16.23, 18.49, 20.71
22.61, 23.83, 25.63
16.24, 18.48, 20.66
22.59, 23.80
20.69, 22.59, 23.79,
25.09, 25.56, 26.75
25.84, 28.73, 29.67
30.42, 32.14
10.92, 11.53, 12.29,
12.98, 14.55, 15.48
19.03, 20.00, 22.84,
25.67, 25.96, 26.32
28.80, 29.25, 30.83
32.16, 34.40
      than one peak listed does not  Implicate the total number of peaks
characteristic of the multi-component  pesticide.  Listed peaks Indicate  only the
ones chosen for quantification.

-------
Table 3.  Matrix Effect on Single Laboratory Recovery
and Precision of Organohallde Pesticides and Aroclors
Compound
Alachlor
Aldrln




Chlordane



01eldr1n





Endrln





Heptachlor





Matrix1
RU
RU

6U

TU
RU

TU

RU

SW

TU

RU

6U

TU

RU

6U

TU

Number
of
Samples
6
7
3
7
3
3
8
7
8
8
7
7
7
6
8
6
6
6
6
6
7
6
. 6
6
7
6
8
8
Spike
Level
(uq/L)
0.50
0.05
1.8
0.05
1.2
1.2
0.17
3.4
0.17
3.4
0.10
3.6
0.10
3.6
0.10
3.6
0.10
3.6
0.10
3.6
0.10
3.6
0.032
1.2
0.032
1.2
0.032
1.2
Aver«9e
Accuracy
(% Recovery)
102
106
105
86
92
94
HA
NA
104
95
87
114
67
94
92
81
119
99
94
IX
106
85
77
80
37 .
r.
200
106
Relative
Standard
Dev1at1on(X)
13.1
18.9
4.9
18.9
15.7
5.6
8.0
3.6
11.8
10.1
19.6
8.0
15.0
9.1
17.1
17.3
25.0
6.6
21.5
11.3
13.2
14.6
13.3
9.3
18.3
13.8
11.3
15.8

-------
Table 3.  (Continued)
Compound
Heptachlor Epoxlde


Hexachl orobenzene


Llndane


Methoxychlor
Toxaphene

Aroclor 1016

Aroclor 1221

Matrix1
RU
6U
TU
RW
6W
TU
RU
GW
TU
RU
RU
TU
RU
TU
RU
TU
Number Spike Average
of Level Accuracy
Samples (ng/t) « Recovery)
8
8
7
6
6
5
7
8
7
7
7
6
7
7
7
7
8
6
5
6
8
8
• 8
8
8
8
7
7
0.04
1.4
0.04
1.4
0.04
1.4
0.0025
0.09
0.002
0.09
0.0027
0.09
0.03
1.2
0.03
1.2
0.03
1.2
2.10
7.03
10
80
10
80
1.0
1.0
•180
. *
180
100
115
90
103
112
81
104
103
91
101
in
88
91
111
88
109
103
93
100
98
HA
HA
110
114
HA
97
HA
92
Relative
Standard
.Dev1at1on(X)
15.6
5.7
15.8
6.7
6.7
7.3
13.0
6.4
12.0
4.4
15.6
15.2
7.1
4.5
8.8
3.1
7.9
19.8
21.0
11.1
12.6
15.3
8.6
11.8
6.6
7.7
8.3
10.4

-------
                            Table 3.  (Continued)
Compound
Aroclor 1232

Aroclor 1242

Aroclor 1248
Aroclor 1254

Aroclor 1260

Matrix1
RW
TW
RW
TU
RW
TW
RW
TW
RW
TW
Number
of
Samples
8
8
6
7
7
8
8
8
7
6
Spue
Level
(uq/l)
3.9
4.3
4.7
4.8
3.6
3.4
1.8
1.7
2.0
1.8
Average
Accuracy
(t Recovery)
NA
8$
NA
96
NA
84
NA
85
NA
88
Relative
Standard
Dev1at1on(X)
13.5
8.5
6.0
7.7
11.5
11.8
10.4
13.9
20.7
18.0
Matrix Identities

   RW • Reagent water
   GW • Ground water
   TW - Tap Water

   NA «  Not applicable.  A separate set of aqueous standards was not
         analyzed, and the values shown for RW were used to calculate X
         recovery for the TW matrix.

-------
























r ,
a
o
*» 11
tj ° *d
• 3 '£ 8. -3
•2 S 12 S-5 •" £
*f^ *• CV DU *T^ *\j
J-d •-« H _H M — '
C q O TJ O V
•H fl »< •-! »H i-J -H
X 3 ft
0
O
















I











i
—• V
1 !
£ 3
£ o
a
J^
j
u
§, 3
•H a
^
c
•M
O
i
o










t>
ra



S
1
2







[













V,





































J











3
J
M




-




,— .**














•i
i
j



1 1 1 1 1 1 1 1 1 1
4 • • 10 12 14 16 IB 20 22

(
« COLUMN: Fused silica capillary
> . LIQUID PHASE: DB-1
| FILM THICKNESS: 1.0wm
3 COLUMN DIMENSIONS: 0.32nw ID, 30 M Ion














1
3




•5
|^
•«
&



I °
fl|
ll



















t



. 	 A. 	
i ii i i ii ii I J —
24 » 28 30 32 34 30 38 40 42 44
                                          TIME (MIN)
Figure 1.  Hexane spiked at 7.71  ug/L with heptachlor and llndane; 9.14 ug/L with heptachlor epoxlde;
           11.4 ug/L with aldrln  and hexachlorobenzene; 23 ug/L with butachlor, chlorpyrlfos, chlorpyrlfos-
           methyl. dlclobenll, dleldrln, endrln, metolochlor, and propachlort  and 44.9 ug/L with
           mthoxychlor.

-------
                                                    	A
                                                                   COLUMN:   Fused silica capillary
                                                                   LIQUID PHASE;   OB-1
                                                                   FILM THICKNESS:   1.0Mn
                                                                   COLUMN DIMENSIONS:  0.32tm ID, 30 M long
     a    4   e    •   10   la  14   ie   is   ao  aa   24   as   as   so  sa   34   M   as  40   42   44
                                           TIME (MIN)
Figure 2.   Hexane spiked it 11.4 ug/L with Arodor 1016.

-------
                                                                    COLUMN: Fused silica capillary
                                                                    LIQUID PHASE:  OB-1
                                                                    FILM THICKNESS:  1.0pm
                                                                    COLUMN DIMENSIONS:  0.32nw ID. 30 M long
                                                                                    .A/V	
    a    4    •    •   10  la  14   10   10  20  22   94   2ft   88   30  39   94   30   38  40   42   44
                                          TINE (MIN)

Figure 3.  Hexane spiked  at 171.4 ug/L with Aroclor 1221.

-------
                                                                   COLUMN:  Fused silica capillary
                                                                   LIQUID PHASE:  DB-1
                                                                   FILM THICKNESS:  1.0p»
                                                                   COLUMN DIMENSIONS:  0.32wn ID, 30 M long
                                                                      	/\	^
                       10   12  14   18   18  20  22   24   20   28   90   32  34   38   38  40   42   44
                                          TIME (MIN)
Figure 4.   Hexane spiked  at  57.1 ug/L with Aroclor 1232.

-------

 Jl
                                                              COLUMN:  Fused silica capillary
                                                              LIQUID PHASE:  OB-1
                                                              FILM THICKNESS:  1.0pm
                                                              COLUMN DIMENSIONS)  0.32mm ID, 30 M long
                                                                   I
    a    4   6   8    10  la  14  16   16   20  22  24   »  28   30  32  34   M  38   40  42   44
                                        TIME (MIN)
                                    •

Figure  5.  Hexane spiked at 57.1 ug/L with Aroclor 1242.

-------
                                                                    COLUMN:  Fused silica capillary
                                                                    LIQUID PHASE:  DB-1
                                                                    FILM THICKNESS:  1.0pm
                                                                    COLUMN DIMENSIONS:  0.32im ID, 30 M long
10  12   14  19
IS   20  22   24
 TIME (HIM)
                                                                 28   30   32   34  M   38  4O  42   44
Figure 6.  Hexane spiked at 57.1  ug/L with Aroclor 12*8.

-------
                                                                    COLUMN:  Fused siHca capillary
                                                                    LIQUID PHASE:  OB-1
                                                                    FILM THICKNESS:  LOpm
                                                                    COLUMN DIMENSIONS:   0.32m ID. 30 H long
    a    4    •   8    10  la   14  16   18   20  22   24   2ft  28   30   32  34   36   38  40   42   44
                                          TIME (MIN)
Figure 7.  Hexane spiked at 42.9 ug/L with Aroclor 1254 .

-------
                                                            COLUMN:  Fused silica capillary
                                                            LIQUID PHASE:  OB-1
                                                            FILM THICKNESS:  1.0|im
                                                            COLUMN DIMENSIONS:  0.32mm ID, 30 M long
 JU
r   .   .
A..
                                                 i	i_
                    10  la  14
  la   20   22
   TINE (NIN)
24  29  28  30   32   34  36   38  40  42   44
Figure 8.  Hexane spiked at 34.3 ug/L with Aroclor 1260.

-------
                                                                  COLUMN:  Fused silica capillary
                                                                  LIQUID PHASE:  DB-1
                                                                  FILM THICKNESS:  1.0pm
                                                                  COLUMN DIMENSIONS:  0.32nm ID,  30 M long
                       10   12   14  16   18   20  22   24   29  28   30   32   94   36   36  40  42   44
                                         TIME (MIN)
Figure 9.  Hexane spiked at 28.6 ug/L with chlordane.

-------