-------

-------
       EXHIBIT A
SUMMARY OF REQUIREMENTS
                                          OLM01.0

-------
                                   SECTION  I
                             GENERAL REQUIREMENTS
       The  Contractor shall  use  proven  instruments and  techniques  to  identify
 and measure  the  concentrations  of volatile, semivolatile, and pesticide
 compounds  listed on the Target  Compound List  (TCL)  in  Exhibit C.  The
 Contractor shall employ state-of-the-art GC/MS and/or  GC/EC procedures to
 perform all  analyses,  including the necessary preparations for analysis.

       In Exhibit D,  the EPA provides the Contractor with the specific
 analytical procedures  to be used and defines the specific application of
 these  procedures to this contract.  For volatiles and  semivolatiles, this
 includes instructions  for sample preparation, gas chromatographic screening,
 mass spectrometric  identification, and data evaluation.  Specific ions used
 for searching  the mass spectral  data for each compound are included.  For
 pesticides,  this includes instructions for sample preparation, gas
 chromatography,  confirmation of  identification by gas  chromatography and/or
 mass spectrometry,  and data evaluation.

       The  Contractor shall  prepare extracts and dilutions of samples.  The
 Contractor shall screen extracts by methods of his  choice (soil
 characterization mandatory; water characterization  optional) at an initial
 extract concentration.  Then, based on the screening response, the Contractor
 shall  use  the  specific analytical methods described in Exhibit D  to extract
 and concentrate  samples to  achieve the Contract Required Quantitation Limits
 (CRQL) listed  in Exhibit C.  Exhibit D lists the analytical methods and
 starting points  to be utilized for each of the target  compounds.

       During preparation,  the Contractor shall fortify all seraivola.tile and
 pesticide  samples, blanks, matrix spikes,  and matrix spike duplicates with
 the surrogate  spiking compounds listed in Exhibit D.   The Contractor shall
 fortify all volatile samples, blanks,  matrix spikes, and matrix spike
duplicates with  the system monitoring compounds listed in Exhibit 0.
Additionally,  all sample semivolatile extracts and  aliquots for volatile
organics analysis shall be  spiked with the internal standard compounds listed
 in  Exhibit D before injection or purging.

      Additionally, for each sample analyzed by GC/MS, the Contractor shall
conduct mass spectral library searches to determine the possible  identity of
up  to  ten  (10) non-system monitoring compounds volatile components and up to
twenty (20) non-surrogate semivolatile components that are not on the Target
Compound List  (Exhibit C).

      Exhibit  F  contains chain-of-custody and sample documentation
requirements which  the Contractor must follow in processing samples under
this contract, and specifies requirements for written  laboratory  standard
operating procedures.

      Sample analysis data, sample documentation and other deliverables shall
be  reported as specified in Exhibit B.   Specifications for reporting data in
computer-readable form appear in Exhibit H.

                                   A-2                                OLM01.0

-------
      To ensure proper understanding of language utilized in this contract,
Exhibit G contains a glossary of terms.  When a term is used in the text
without explanation, the glossary meaning shall be applicable.

      The samples to be analyzed by the Contractor are from known or
suspected hazardous waste sites and, potentially,  may contain hazardous
organic and/or inorganic materials at high concentration levels.  The
Contractor should be aware of the potential hazards associated with the
handling and analyses of these samples.  It is the Contractor's
responsibility to take all necessary measures to ensure the health and safety
of its employees.

      In addition, the Contractor must be aware of the importance of
maintaining the integrity of the data generated under the contracts as it is
used to make major decisions regarding public health and environmental
welfare.  In addition, it may be used in litigation against potentially
responsible parties in the enforcement of Superfund legislation.

      Prior to accepting any samples from the Agency, the Contractor shall
have, in-house, the appropriate standards for all target compounds listed in
Exhibit C.
                                   A-3                                OLM01.0

-------
                                 SECTION II
                           SUMMARY OF REQUIREMENTS


I.    For each sample,  the Contractor  shall  perform  the  following  tasks:

     A.   Task I  Receive and Prepare Hazardous Waste Samples.

         1.   Receive and handle samples under the chain-of-custody
              procedures described in Exhibit F.

         2.   Prepare samples as described in Exhibit D.  VOA analysis of
              water or soil samples must be completed within 10 days of VTSR
               (Validated Time of Sample Receipt).   Separatory funnel
              extractions for pesticides in water samples must be completed
              within 5 days of VTSR.  Sonication extractions for pesticides
              and/or semivolatiles in soil samples must be completed within
              10 days of VTSR.  Continuous liquid-liquid extraction for
              semivolatile samples must be started within 5 days of VTSR.

              Extracts of either water or soil samples must be analyzed
              within 40  days of extraction.  This does not release the
              Contractor from the data turnaround time specified in Exhibit
              B, Section I.

     B.   Task II  Analysis for Identification of Specific Organic Compounds.

         1.   Extracts and aliquots prepared in Task I shall be analyzed by
              GC/EC and  GC/MS techniques given in Exhibit D for the target
              compounds  listed in Exhibit C.

         2.   The target compounds listed in Exhibit C shall be identified
              as described in the methodologies given in Exhibit D.
              Automated  computer programs may be used to facilitate the
              identification.

     C.   Task III  Qualitative Verification of the Compounds Identified  in
         Task II.

         1.   The volatile and semivolatile compounds analyzed by GC/MS
              techniques and initially identified in Task II shall be
              verified by an analyst competent in the interpretation of mass
              spectra by comparison of the suspect mass spectrum to the mass
              spectrum of a standard of the suspected compound.  This
              procedure  requires the use of multiple internal standards.
              Two criteria must be satisfied to verify the identifications:

              a.   Elution of the sample component at the same GC relative
                   retention time as the standard component.

              b.   Correspondence of the sample component and standard
                   component mass spectra.

                                   A-4                        OLM01.4    3/91

-------
     For establishing correspondence of the GC relative retention
     time (RRT),  the sample component RRT must compare within +
     0.06 RRT units of the RRT of the standard component.   For
     reference, the calibration standard must be run on the same
     12-hour time period as the sample.

     For comparison of standard and sample component mass  spectra,
     mass spectra obtained on the Contractor's GC/MS are required.
     Once obtained, these standard spectra may be used for
     identification purposes only if the Contractor's GC/MS meets
     the DFTPP or BFB daily instrument performance check
     requirements specified in Exhibit D.  The standard spectra
     used may be from a laboratory generated library on the same
     instrument or obtained from the calibration standard  run used
     to obtain reference RRTs.   The requirements for qualitative
     verification by comparison of mass spectra are as follows:

     a.   All ions present in the standard mass  spectrum at a
          relative intensity greater than 10 percent (most abundant
          ion in the spectrum equals 100 percent)  must be  present
          in the  sample spectrum.

     b.   The relative intensities of ions specified above must
          agree within + 20 percent between the  standard and sample
          spectra.
     c.   Ions greater than 10  percent  in the  sample  spectrum but
          not present in the standard spectrum must be  considered
          and accounted for by  the  analyst making the comparison.
          When GC/MS computer data  processing  programs  are  used to
          obtain the sample component spectrum,  both  the  processed
          and the raw spectra must  be evaluated.  In Task  III,  the
          verification process  should favor false positives.

3.   If a compound analyzed by  GC/MS techniques and initially
     identified in Task II cannot be verified by all  of the
     criteria in items 1 and 2  above, but in the technical
     judgement of the mass spectral interpretation specialist the
     identification is correct, then the Contractor shall report
     that identification, and proceed with quantification in Task
     IV.

4.   The Pesticide/Aroclor compounds listed in Exhibit  C and
     analyzed by GC/EC techniques shall have their identifications
     verified by an analyst competent in the interpretation of gas
     chromatograms.   Two criteria must  be satisfied to  verify the
     identifications:

a.   Elution of the sample component within the retention time
     window (established by the procedures in Exhibit D)  of the
     standard component analyzed on the same GC column  and
     instrument, as specified in Exhibit D.
                          A-5                               OLM01.0

-------
     b.   Analysis  of the  sample  and  standard on a second GC column with
          a stationary phase  with retention characteristics dissimilar
          to that used in  a.  above, and meeting the  same criteria  for
          elution of the sample component  and the standard as  in a.
          above.

D.   Task IV  Quantification  of Compounds  Verified  in Task  III.

     1.   The Contractor shall quantify components analyzed by GC/MS
          techniques,  identified  in Task -II and verified in Task III by
          the internal standard method stipulated in Exhibit D.  Where
          multiple  internal standards are  required by EPA, the
          Contractor shall perfb'rm quantitation utilizing the  internal
          standards specified in  Exhibit D.

     2.   The Contractor shall determine response factors for  each
          12-hour time period of  GC/MS analysis and  shall include  a
          calibration check of the initial five point calibration  as
          described in Exhibit D.

     3.   The Contractor shall quantify components analyzed by GC/EC
          techniques,  identified  in Task II and verified in Task III by
          the external standard method stipulated in Exhibit D.

     4.   The Contractor shall perform an  initial three-point
          calibration,  verify its  linearity, determine the breakdown of
          labile components,  and  determine calibration factors for all
          standards analyzed  by GC/EC techniques as  described  in Exhibit
          D.

E.   Task V  Tentative Identification of Non-target  Sample  Components.

     1.   For each  analysis of a  sample, the Contractor shall  conduct
          mass spectral library searches to determine tentative compound
          identifications  as  follows. For  each volatile fraction,  the
          Contractor shall conduct a  search to determine the possible
          identity  of up to ten (10)  organic compounds of greatest
          concentration which are  not system monitoring compounds  and
          are not listed in Exhibit C.  For each semivolatile  fraction,
          Che Contractor shall conduct a search to determine the
          possible  identification of  up to twenty (20) non-surrogate
          organic compounds of greatest concentration which are not
          listed in Exhibit C.  In performing searches, the most recent
          release of the NIST/EPA/MSDC mass spectral library must  be
          used.  NOTE:  Substances with responses less than 10 percent
          of the nearest internal  standard are not  required to be
          searched  in this fashion.
                               A-6                          OLM01.3  2/91

-------
     2.    Only  after visual comparison of sample spectra with the
          spectra  from  the library searches will the mass spectral
          interpretation specialist assign a tentative identification.
          If  the compound does not meet  the identification criteria of
          Task  III, it  shall be  reported as unknown. The mass spectral
          specialist should give additional classification of the
          unknown  compound, if possible  (i.e., unknown aromatic, unknown
          hydrocarbon,  unknown acid type, unknown chlorinated compound).
          If  probable molecular  weights  can be distinguished, include
          them.

     3.    The Contractor shall not report as semivolatile tentatively
          identified compounds (TIC) any target compounds from  the
          volatile fraction (i.e., do not report late eluting volatile
          compounds as  TICs in the semi-volatile analysis).  However, the
          Contractor may report  pesticide target compounds that appear
          as  semivolatile tentatively identified compounds.

F.    Task VI  Quality Assurance/Quality  Control Procedures.

     1.    All specific  quality assurance procedures prescribed  in
          Exhibits D and E shall be strictly adhered to by the
          Contractor.   Records documenting the use of the protocol shall
          be  maintained in accordance with the document control
          procedures prescribed  in Exhibit F, and shall be reported in
          accordance with Exhibit B, Reporting Requirements and
          Deliverables, and Exhibit H, Data Dictionary and Format for
          Data  Deliverables in Computer-Readable Format.

     2.    The Contractor shall establish a Quality Assurance Plan (QAP)
          with  the objective of  providing sound analytical chemical
          measurements.  This program shall incorporate the quality
          control  procedures, any necessary corrective action,  and all
          documentation required during  data collection as well as the
          quality  assessment measures performed by management to ensure
          acceptable data production.

     3.    The Contractor shall perform one spiked sample analysis
          (matrix  spike) and one duplicate spiked sample analysis
          (matrix  spike duplicate) for each group of samples of a
          similar  matrix (for water or soil samples) and concentration
          level (for volatile and semivolatile soil samples only) for
          the following, whichever is most frequent:

          o     Each Case of field samples received, OR

          o     Each 20  samples in a Case, OR

          o     Each 14  calendar  day period during which field samples in
               a Case were received (7 calendar day period for  14-day
               data turnaround contracts) (said period beginning with
               the receipt of the first  sample in that Sample Delivery
               Group).
                               A-7                          OLM01.2  1/91

-------
     Matrix spikes and matrix spike  duplicates  shall  be  carried
     through the entire analytical process  from extraction  to  final
     GC/MS or GC/EC analysis,  including all Contract
     Performance/Delivery Requirements  (see Contract  Schedule).

4.   The Contractor shall prepare and analyze one  laboratory
     reagent blank (method blank) for ea.ch  group of samples of a
     similar matrix (for water or soil  samples), extracted  by  a
     similar method (separatory funnel,  continuous liquid-liquid
     extraction, or sonication,  as specified in Exhibit  D),  and  a
     similar concentration level (for volatile  and semivolatile
     soil samples only) for the following,  whichever  is  most
     frequent:

     o    Each  Case  of field samples  received,  OR

     o    Each  20 samples in a Case,  including  matrix spikes and
          reanalyses,  OR

     o    Each  14 calendar day period (7 calendar  day period for
          14-day data  turnaround contracts)  during which field
          samples in a Case were received (said period beginning
          with  the receipt of the first  sample  in  that Sample
          Delivery Group),  OR

     o    Whenever samples are extracted.

     Volatile analysis requires one  method  blank for  each 12-hour
     time period when  volatile target compounds are analyzed.

     Semivolatile and  pesticide method  blanks shall be  carried
     through the entire analytical process  from extraction  to  final
     GC/MS or GC/EC analysis,  including all Contract
     Performance/Delivery Requirements  (see Contract  Schedule).

5.   The Contractor shall verify instrument performance  for each
     12-hour time period, to include the following:
     Decafluorotriphenylphosphine  (DFTPP) and/or Bromofluorobenzene
     (BFB) as applicable, and a specific calibration  using
     standards  of defined concentration to  monitor response,
     retention time, and mass spectra.

     Additional quality control shall be conducted in the form of
     the analysis of laboratory evaluation  samples submitted to  the
     laboratory by the Agency.  The  results of  all such  control  or
     laboratory evaluation samples may  be used  as  grounds for
     termination of noncompliant Contractors.   "Compliant
     performance" is defined as that which  yields  correct compound
     identification and concentration values as determined  by  the
     Agency,  as well as meeting the  contract requirements for
     analysis (Exhibit D), quality assurance/quality  control
     (Exhibit E), data reporting and other  deliverables  (Exhibits B
     and H),  and sample custody, sample documentation and SOP
     documentation (Exhibit F).

                          A-8            •             OLM01.2   1/91

-------
B.   The EPA has provided to the Contractor formats  for the  reporting of data
     (Exhibits B and H).   The Contractor shall  be  responsible  for completing
     and returning analysis data sheets  and submitting computer-readable data
     on diskette in the  format specified in this SOW and within the  time
     specified in the Contract Performance/Delivery  Schedule.

     1.    Use of formats  other than  those designated by EPA  will  be  deemed  as
          noncompliance.   Such data  are  unacceptable.   Resubmission  in the
          specified format at no additional  cost to  the government will be
          required.

     2.    Computer generated forms may be submitted  in the hardcopy  data
          package(s)  provided that the forms are in  EXACT EPA  FORMAT.   This
          means  that  the  order of data elements is the same  as  on each EPA
          required form,  including form  numbers and  titles,  page  numbers and
          header information.

     3.    The data reported by the Contractor on the hardcopy  data forms and
          the associated  computer-readable  data submitted by the  Contractor
          must contain identical information.   If during government
          inspection  discrepancies are found, the Contractor shall be
          required to resubmit either the hardcopy forms, or the  computer
          readable data,  or both sets of data at no  additional  cost  to the
          government.

C.   The Contractor shall provide analytical equipment and  technical
     expertise for this  contract as  specified by the following:

     1.    The Contractor  shall have  sufficient  gas chromatograph/electron
          capture/data systems (GC/EC/DS) and gas chromatograph/mass
          spectrometer/data system (GC/MS/DS) capability to  meet  all the
          terms  and conditions of the Contract.  The Contractor shall
          maintain, at a  minimum, all analytical equipment allocated for this
          contract at the time of contract  award.

     2.    The Contractor's instrument systems shall  have the following:

          a.   The GC/MS  shall be equipped with a glass  jet  separator when
              using packed  columns.

          b.   The computer  shall be interfaced by hardware  to  the mass
              spectrometer  and be capable of acquiring  continuous mass  scans
              for  the duration of the chromatographic  program.

          c.   The computer  shall be equipped with mass  storage devices  for
              saving  all  data  from  the  GC/MS runs.

          d.   Computer software  shall be available  to  allow searching GC/MS
              runs for specific  ions and plotting the  intensity  of  the  ions
              with respect  to  time  or scan  number.
                                   A-9                         OLM01.2   1/91

-------
     e.   A computer data system must be interfaced to  the  mass
          spectrometer that allows the continuous  acquisition and
          storage,  on machine readable media,  of all mass spectra
          obtained throughout the duration of the  chromatographic
          program.   The computer must have software that allows
          searching any GC/MS data file for ions of a specified  mass  and
          plotting such ion abundances versus time or scan  number.  This
          type of plot is defined as  an Extracted  Ion Current Profile
          (EICP).   Software must also be available that allows
          integrating the abundance in any EICP between specified time
          or scan number limits.  Also,  for the non-target  compounds,
          software must be available  that allows for the comparison of
          sample  spectra against reference library spectra.   The most
          recent  release of the NIST/EPA/MSDC mass spectral library
          shall be used as the reference library.   The  data system must
          be capable of flagging all  data files that have been edited
          manually by laboratory personnel.

     f.   The GC/MS shall be equipped with a GC to MS interface  capable
          of extending a fused silica capillary column  into the  ion
          source.   The column is to be 30 meters long by 0.25 or 0.32 mm
          inside  diameter,  bonded DB-5,  fused silica, or equivalent.

     g.   The GC/EC/DS for pesticide  analysis shall be  equipped  with
          wide bore capillary columns and a suitable detector and data
          system  as described in Exhibit D.

3.   The Contractor shall use a magnetic tape storage device capable  of
     recording data and suitable for  long-term, off-line storage.  The
     Contractor shall retain all raw  GC/MS data acquired under this
     contract on  magnetic tape in appropriate instrument manufacturer's
     format.  The Contractor is required to retain the  magnetic  tapes
     with associated hardcopy tape logbook identifying  tape contents
     (see Exhibit E) for 365 days after data submission.  During that
     time, the Contractor shall submit tapes and logbook within  7 days
     of request from EMSL/LV or the Administrative Project  Officer
     (APO),  as specified in the Contract Performance/Delivery Schedule
     and Exhibit  E.

4.   The Contractor shall have a computerized MS library search  system
     capable of providing a forward comparison, utilizing the standard
     spectra contained in the mass spectral library.  The  most recent
     release of the NIST/EPA/MSDC mass spectral library, must be used.

     a.   The system shall provide a  numerical ranking  of  the standard
          spectra most closely corresponding to the sample  spectra
          examined.

     b.   The data  system shall have  software capable of removing
          background signals from spectra.

5.   The Contractor shall have, in-house and operable,  a device  capable
     of analyzing purgeable organics  as described  in Exhibit D.
                              A-10                         OLM01.3  2/91

-------
6.   The Contractor shall have, in-house,  the appropriate standards for
     all target compounds listed in Exhibit C prior to accepting any
     samples from the Agency.

The Contractor shall have an IBM or IBM-compatible mini-computer or PC
capable of recording required sample data on 5.25 inch floppy
double-sided double-density 360 K-byte or 1.2 M-byte or a 3.5 inch
double-sided, double density 720 K-byte or 1.44 M-byte diskette, in
ASCII text file format, and in accordance with the file, record and
field specifications listed in Exhibit H.

The Contractor shall designate and utilize key personnel to perform the
minimum functional requirements necessary to meet the terms and
conditions of this contract.  The EPA reserves the right to review
personnel qualifications and experience.  The minimum functional
requirements are listed below:

o    Project Manager

o    GC/MS Laboratory Supervisor

o    GC/EC Laboratory Supervisor

o    Sample Preparation Laboratory Supervisor

o    Quality Assurance Officer

o    Systems Manager

o    Programmer Analyst

o    GC/MS Operator

o    Mass Spectral Interpretation Specialist

o    GC/EC Operator

o    Pesticide Residue Analysis Expert

o    Extraction/Concentration Expert

o    Sample Custodian

o    Data Reporting and Delivery Officer

The Contractor shall respond within seven days to written requests  from
data recipients for additional information or explanations that result
from the Government's inspection activities unless otherwise specified
in the contract.
                              A-ll                         OLM01.2  1/91

-------
G.   The Contractor shall store all samples and unused sample  volume  at 4°C
     (±2'C), protected from light,  from time of receipt until  60 days after
     data submission.  Samples and unused sample volumes must  be stored
     separately from sample extracts and standards.   The Contractor shall
     preserve all sample extracts after analysis in bottles/vials with
     Teflon-lined septa and shall maintain stored extracts at  4"C (±2°C).
     The Contractor is required to retain the sample extracts  for 365 days
     after data submission.  During that time,  the Contractor  shall submit
     the extracts within seven days of a written request by TPO/APO,  as
     specified in the Contract Performance/Delivery Schedule.

H.   The Contractor shall adhere to chain-of-custody procedures described in
     Exhibit F.  Documentation, as described therein, shall be required to
     show that all procedures are being strictly followed.  This
     documentation shall be reported as the complete SDG file  purge (see
     Exhibit B).

I.   Sample shipments to the Contractor's facility will be scheduled and
     coordinated by the EPA CLP Sample Management Office (SMO) acting on
     behalf of the Administrative Project Officer.  The Contractor shall
     communicate with SMO personnel by telephone, as necessary throughout the
     process of sample scheduling,  shipment, analysis and data reporting, to
     ensure that samples are properly processed.

     If there are problems with the samples (e.g., mixed media, containers
     broken or leaking) or sample documentation/paperwork (e.g., Traffic
     Reports not with shipment, sample and Traffic Report numbers do not
     correspond)  the Contractor shall immediately contact SMO  for resolution.
     The Contractor shall immediately notify SMO regarding any problems and
     laboratory conditions that affect the timeliness of analyses and data
     reporting.  In particular, the Contractor shall notify SMO personnel in
     advance regarding sample data that will be delivered late and shall
     specify the  estimated delivery date.

J.   Sample analyses will be scheduled by groups of samples, each defined as
     a Case and identified by a unique EPA Case number assigned by SMO. A
     Case signifies a group of samples collected at one site or geographical
     area over a  finite time period, and will include one or more field
     samples with associated blanks.  Samples may be shipped to the
     Contractor in a single shipment or multiple shipments over a period of
     time,  depending on the size of the Case.

     A Case consists of one or more Sample Delivery Group(s).   A Sample
     Delivery Group (SDG) is defined by the following, whichever is most
     frequent:

     o   Each Case  of  field samples  received, OR

     o   Each 20  field samples  within a  Case, OR

     o   Each 14  calendar  day period (7  calendar day period for 14-day data
         turnaround contracts)  during which field samples  in a Case are
         received (said period  beginning with the receipt  of the first sample
         in  the Sample  Delivery Group).
                                   A-12                         OLM01.2  1/91

-------
     Samples may be assigned to Sample Delivery Groups  by matrix (i.e.,  all
     soils in one SDG,  all waters in another),  at the discretion of the
     laboratory.  Such assignment must be made  at the time the samples  are
     received,  and may not be made retroactively.

     All data (hardcopy and computer-readable format) for all samples in a
     Sample Delivery Group are due concurrently to. all  data recipients  as
     stipulated in the Delivery Schedule in Exhibit B,  Section I.   Data for
     all samples in a Sample Delivery Group must be submitted together  (in
     one package) in the order specified in Exhibit B.   The Sample Delivery
     Group number is the EPA sample number of the first sample received in
     the SDG.  When several samples are received together in the first  SDG
     shipment,  the SDG number shall be the lowest sample number (considering
     both alpha and numeric designations) in the first  group of samples
     received under the SDG.  The SDG number is reported on all data
     reporting forms.

     The SDG Receipt Date is the day the last sample in the SDG is received.
     Data for all samples in the SDG are due as stipulated in the Delivery
     Schedule in Exhibit B.

     The Contractor is responsible for identifying each Sample Delivery Group
     as samples are received, through proper sample documentation (see
     Exhibit B) and communication with SMO personnel.

K.   Each sample received by the Contractor will be labeled with an EPA
     sample number, and accompanied by a Traffic Report form bearing the
     sample number and descriptive information  regarding the sample.  The
     Contractor shall complete and sign the Traffic Report, recording the
     date of sample receipt and sample condition on the receipt for each
     sample container.

     The Contractor shall submit signed copies  of Traffic Reports for all
     samples in a Sample Delivery Group to SMO  within 3 calendar days
     following receipt of the last sample in the Sample Delivery Group.
     Traffic Reports shall be submitted in Sample Delivery Group sets (i.e.,
     all Traffic Reports for a Sample Delivery  Group shall be clipped
     together)  with an SDG Cover Sheet containing information regarding the
     Sample Delivery Group, as specified in Exhibit B.

L.   EPA Case numbers (including SDG numbers) and EPA sample numbers shall be
     used by the Contractor in identifying samples received under this
     contract both verbally and in reports/correspondence.

M.   Samples will be shipped routinely to the Contractor through an overnight
     delivery service.   However, as necessary,  the Contractor shall be
     responsible for any handling or processing required for the receipt of
     sample shipments,  including pick-up of samples at  the nearest servicing
     airport, bus station or other carrier service within the Contractor's
     geographical area.  The Contractor shall be available to receive sample
     shipments at any time the delivery service is operating, including
     Saturdays.
                                   A-13                               OLM01.0

-------
The Contractor shall accept all samples scheduled by SMO,  provided that
the total number of samples received in any calendar month does not
exceed the monthly limitation expressed in the contract.   Should the
Contractor elect to accept additional samples, the Contractor shall
remain bound by all contract requirements for analysis of those samples
accepted.
                              A-14                               OLM01.0

-------
                                 SECTION III
                 DETAILED TECHNICAL & MANAGEMENT REQUIREMENTS
As cited in Section II the Contractor shall have the following technical and
management capabilities.   Note:  For those technical functions which require
a minimum educational degree and experience, an advanced degree in chemistry
or any scientific/engineering discipline, (e.g., Master's or Doctorate) does
not substitute for the minimum experience requirements.

Any personnel changes affecting the key personnel as stated in Exhibit A,
Section III, Items I and II, the Contractor shall notify in writing the
Technical Project Officer and the Administrative Project Officer within 14
days of the personnel change.  The Contractor shall provide a detailed resume
to the Technical Project Officer, Administrative Project Officer, and EMSL/LV
for the replacement personnel within 14 days of the Contractor's assignment
of the personnel.  The resume shall include position description of titles,
education (pertinent to this contract),  number of years of experience
(pertinent to this contract) month and year hired, previous experience and
publications.

I.   TECHNICAL CAPABILITY

     A.    Technical  Functions

          1.   GC/MS Laboratory Supervisor

              a.   Responsible for all technical efforts of the GC/MS
                   laboratory to meet all terms and conditions of the EPA
                   contract.

              b.   Qualifications:

                   (1)  Education:

                         Minimum of Bachelor's degree in chemistry or any
                         scientific/engineering discipline.

                   (2)  Experience:

                         Minimum of three years of laboratory experience  in
                         operating a GC/MS, including at least one year  of
                         supervisory experience.

          2.   GC/EC Laboratory Supervisor

              a.   Responsible for all technical efforts of the GC/EC
                   laboratory to meet all terms and conditions of the EPA
                   contract.
                                   A-15                         OLM01.2  1/91

-------
     b.   Qualifications:

          (1)   Education:
                Minimum of Bachelor's  degree  in chemistry  or  any
                scientific/engineering discipline.
          (2)   Experience:

                Minimum of three years  of laboratory experience  in
                operating a GC/EC,  including at  least one  year of
                supervisory experience.

3.   Sample Preparation Laboratory Supervisor

     a.   Responsible for all  technical  efforts  of sample
          preparations to meet all  terms and conditions  of the EPA
          contract.

     b.   Qualifications:

          (1)   Education:

                Minimum of Bachelor's degree in  chemistry  or  any
                scientific/engineering  discipline.

          (2)   Experience:

                Minimum of three years  of laboratory experience  in
                organic sample preparation,  including at least one
                year  of supervisory experience.

4.   Quality Assurance Officer

     a.   Responsible for overseeing the quality assurance aspects
          of the data and reporting directly to  upper management to
          meet all  terms and conditions  of the EPA contract.

     b.   Qualifications:

          (1)   Education:

                Minimum of Bachelor's degree in  chemistry  or  any
                scientific/engineering  discipline.

          (2)   Experience:

                Minimum of three years  of laboratory experience,
                including at least  one  year of applied experience
                with  QA principles  and  practices in an analytical
                laboratory.
                         A-16                         OLM01.2  1/91

-------
5.   Systems Manager

     a.    Responsible  for  the management and quality control of all
          computing  systems  (hardware, software, documentation and
          procedures),  generating, updating, and quality
          controlling  automated  deliverables to meet all  terms and
          conditions of the  EPA  contract,

     b.    Qualifications:

          (1)  Education:
                Minimum of Bachelor's  degree with  four or more
                intermediate courses in programming, information
                management,  database management systems,  or systems
                requirements analysis.

          (2)  Experience:

                Minimum of three years experience  in data or
                systems management or  programming  including one
                year of experience with the software being utilized
                for  data management and generation of laboratory
                reports.

6.   Programmer  Analyst

     a.    Responsible  for  the installation, operation and
          maintenance  of software and  programs generating, updating
          and quality  controlling analytical databases and
          automated  deliverables to meet all terms and conditions
          of the EPA contract.

     b.    Qualifications:

          (1)  Education:

                Minimum of Bachelor's  degree with  four or more
                intermediate courses in programming, information
                management,  information systems, database
                management systems, or systems requirements
                analysis.

          (2)  Experience:

                Minimum of two years experience in systems or
                applications programming including one year of
                experience with  the software being utilized for
                data management  and generation of  laboratory
                reports.
                         A-17                         OLM01.2  1/91

-------
7.   GC/MS Operator Qualifications

     One year of experience  in operating and maintaining GC/MS/DS
     and a minimum of a Bachelor's  degree in chemistry or a
     scientific/engineering  discipline,  or in lieu of the minimum
     education requirement,  three years  of experience in operating
     and maintaining the GC/MS and  interpreting GC/MS data.

8.   Mass Spectral Interpretation Specialist Qualifications

     a.    Education:

          o  Minimum of Bachelor's  degree in chemistry or any
             scientific/engineering discipline.

          o  Training course(s) in mass spectral interpretation.

     b.    Experience:

                Minimum of two years of  experience  in mass  spectral
                interpretation.

9.   GC/EC Operator Qualifications

     One year of experience  in operating and maintaining GC/EC and
     a minimum of a Bachelor's degree  in chemistry or a
     scientific/engineering  discipline,  or in lieu of the minimum
     education requirement,  three years  of experience in operating
     and maintaining the GC/EC and  interpreting GC/EC data.

10.  Pesticide Residue Analysis Expert Qualifications

     a.    Education:
                Minimum of Bachelor's  degree in chemistry or  any
                scientific/engineering discipline.

     b.    Experience:

                Minimum of two years of  experience  in operating and
                maintaining  GC and  interpreting GC  chromatograms.

11.  Extraction/Concentration Expert Qualifications

     a.    Education:

                Minimum of High school diploma  and  a  college  level
                course  in general chemistry.

     b.    Experience:

                Minimum of one year of experience  in
                extraction/concentration.
                         A-18                         OLM01.2  1/91

-------
          12.  Technical Staff Redundancy

              The bidder shall have a minimum of one (1) chemist available
              at any one time as a back-up technical person with the
              following qualifications, to ensure continuous operations to
              accomplish the required work as specified by EPA contract.

              a.   Education:

                         Minimum of Bachelor's degree in chemistry or any
                         scientific/engineering discipline.

              b.   Experience:  Minimum of one year in each of the following
                   areas

                   o   GC  operation and maintenance  for  volatiles  and
                       semivolatiles analyses.

                   o   Mass  spectral interpretation.

                   o   Extraction.
                   o   Pesticide/ Aroclors analysis.

B.    Facilities

     The adequacy  of the facilities and equipment  is  of  equal  importance as
     the technical  staff to  accomplish  the required  work as  specified by the
     EPA contract.

     1.    Sample Receipt Area

          Adequate,  contamination-free,  well ventilated  work space provided
          with chemical resistant bench  top for receipt  and safe handling  of
          EPA samples.

     2.    Storage Area

          Sufficient refrigerator space  to maintain unused EPA sample volume
          for 60 days after  data submission and sample extracts for  365  days
          after data submission.  Samples must be  stored in an atmosphere
          demonstrated  to be free from  all potential  contaminants.   Volatile
          samples must be stored in a refrigerator used  only for storage of
          volatile  samples from this contract.  Samples,  sample extracts,  and
          standards  must be  stored  separately  to prevent cross contamination.
          Semivolatile and pesticide/Aroclor standards and extracts  must be
          stored separately  from volatile standards  and  extracts.

     3.    Sample Preparation Area

          Adequate,  contamination-free,  well-ventilated  work space provided
          with the  following:

          a.   Benches with  chemical resistant tops,  exhaust hoods.   NOTE:
              Standards must be prepared in a glove  box  or isolated area.
                                   A-19                         OLM01.2   1/91

-------
          b.   Source of discilled or demineralized organic-free water.

          c.   Analytical balance(s) located away from draft and rapid change
               in temperature.

C.   Instrumentation

     At a minimum, the Contractor  shall have  the following instruments
     operative at the time  of the  Preaward Site  Evaluation and committed for
     the full duration of the contract.

     1.   Primary Instrument  Requirements  for up to 200 Samples/Month
     Capacity
Fraction
Volatiles
Semivolatiles
Pesticides/
Aroclors
No. of
Ins trument ( s )
1
2
2
Type of
Instrument
GC/MS/DS with
purge and
trap device
GC/MS/DS
GC/EC/DS with
dual column
          This  equates  to  three GC/MS/DS and  2 GC/EC/DS.
     NOTE:  For 300-400 Samples/Month Capacity,  twice as much instrumentation
     is  needed as is listed in item 1.   For 500-600 Samples/Month Capacity,
     three times as much instrumentation is needed as is listed in item 1.
     For 700-800 Samples/Month Capacity,  four times as much instrumentation
     is  needed as is listed in item 1.
     2.    Secondary Instrument  Requirements  for  up  co  400 Samples/Month
          Capacity

          The  Contractor  shall  have  the  following instruments in place and
          operational, at  any  one  time  as a back-up  system;
               Quantity
Instruments
                  One
                  One
                  One
CC/MS/DS
Purge and Trap Device
GC/EC/DS
          Note:   For over 400 samples/raonch capacity,  cwice as rcuch
          instrumentation is  needed as  listed in icem 2.
                                   A-20
                      OLM01.3

-------
     These instruments must be included in the bidder's inventory of
     equipment along with those in 1,  above.

     In addition, the Contractor shall have an in-house stock of
     instrument parts and circuit boards to ensure continuous operatior
     to meet contract-specified holding and turn-around times.

3.   Instrument Specifications

     Instrument specifications are described in detail in the Statement
     of Work (SOW) in the following Exhibits:

     o   Purge  and trap  device Exhibit D

     o   GC/MS/DS              Exhibit D
     o   GC/EC/DS              Exhibit D
Data Management and Handling

1.   Hardware - Contractor shall have  an IBM or IBM-compatible mini-
     computer or PC capable of recording required sample data on 5.25
     inch double-sided,  double-density 360 K-byte or high density 1.2 M-
     byte diskettes; or 3.5 inch double-sided, double-density 720 K-byte
     or 1.44 M-byte diskettes in ASCII text file format and in
     accordance with the file, record and field specifications listed in
     SOW, Exhibit H.

     Other minimum requirements include:

     o    Hard  disk of at least 20 M-bytes.

     o    Asynchronous,  Hayes-compatible  modem capable of at least 2,400
          baud  transmission speed.   In addition,  MNP level 5
          compatibility  is recommended.

     o    Modem capable  of at least 2,400 baud transmission speed which
          is  compatible  with the  EPA Telecommunications Network.

2.   Software - Software, utilized in generating, updating and quality
     controlling analytical databases  and automated deliverables shall
     have the following additional capabilities:

     o    Editing and updating databases.

     o    QC  of automated deliverables.

     o    Controlled access using user ID and  file password protection.

3.   The Contractor shall also be able to submit reports and data
     packages as specified in the Statement of Work Exhibit B.  To
     complete this task, the Contractor shall  be required to provide
     space,  tables and adequate copy machines  to meet the contract
     requirements.
                              A-21                         OLM01.2  1/91

-------
II.  LABORATORY MANAGEMENT CAPABILITY

     The Contractor must have an organization with well-defined
     responsibilities for each individual in the management system to
     ensure sufficient resources for the EPA contract(s)and to maintain
     a successful operation.   To establish this capability, the
     Contractor shall designate personnel to carry out the following
     responsibilities for the EPA contract.   Functions include, but are
     not limited to, the following:

A.   Technical Staff

     Responsible for all technical efforts for the EPA contract.  The
     Contractor shall have adequate number of technical personnel to
     meet the requirements of this contract.

B.   Project Manager

     Responsible for overall  aspects of EPA contract(s) (from sample
     receipt through data delivery)  and shall be the primary contact for
     EPA Headquarters Administrative Project Officer and Regional
     Technical Project Officers.

C.   Sample Custodian

     Responsible for receiving the EPA samples (logging,  handling and
     storage).

D.   Quality Assurance Officer

     Responsible for overseeing the quality assurance aspects of the
     data and reporting directly to upper management.

E.   Document Control Officer

     Responsible for all aspects of data deliverables:  organization,
     packaging,  copying, and  delivery.  Responsible for ensuring that
     all documents generated  are placed in the Complete SDG File for
     inventory and are delivered to the appropriate EPA Regional
     personnel or other receiver.
                              A-22                         OLM01.3  2/91

-------
               EXHIBIT B
REPORTING AND DELIVERABLES REQUIREMENTS
                B-l                         OLM01.0

-------
                              Table of Contents
                                                                         Page
SECTION I:     Contract Reports/Deliverables Distribution 	 B-3

SECTION II:    Report Descriptions and Order of Data
              Deliverables 	 B-6

SECTION III:   Forms Instructions 	 B-24

SECTION IV:    Data Reporting Forms 	 B-53
                                   B-2                                OLM01.0

-------
                                     SECTION I

                     CONTRACT REPORTS/DELIVERABLES DISTRIBUTION
                          (For  35-Day Turnaround Contracts)
The following table reiterates the Contract reporting and deliverables requirements
specified in the Contract Schedule and specifies the distribution that is required
for each deliverable.  NOTE: Specific recipient names and addresses are subject to
change during the term of the contract.  The Administrative Project Officer (APO)
will notify the Contractor in writing of such changes when they occur.

Item
No.
Copies
Delivery
Schedule
Distribution
m (21 (3) C41)
*****A. Standard Operating
        Procedures
60 days after
contract award,
and as required
in Exhibit E.
                        X    X
    *B. Sample Traffic
        Reports
  ***C. Sample Data Summary
        Package
  ***D. Sample Data Package
  ***E. Complete SDG File      1
 ****
*****F.  Qua1i ty
        Assurance
        Plan
  ***G. Data in Computer-
        Readable Form
3 days after
receipt of last
sample in Sample
Delivery Group
(SDG).**

35 days after
receipt of last
sample in SDG.

35 days after
receipt of last
sample in SDG.

35 days after
receipt of last
sample in SDG.

60 days after
contract award,
and as required
in Exhibit E.

35 days after
receipt of last
sample in SDG.
                   X
                        As directed
                   X
                                   B-3A
                           OLM01.3  2/91

-------
                                     SECTION I

                     CONTRACT REPORTS/DELIVERABLES DISTRIBUTION
                          (For  14-Day Turnaround Contracts)
The following table reiterates the Contract reporting and deliverables requirements
specified in the Contract Schedule and specifies the distribution that is required
for each deliverable.  NOTE: Specific recipient names and addresses are subject to
change during the term of the contract.  The Administrative Project Officer will
notify the Contractor in writing of such changes when they occur.
              Item
  No.
Copies
Delivery
Schedule
                    Distribution
                  (1)  (2)  (31
*****A. Standard Operating
        Procedures
60 days after
contract award,
and as required
in Exhibit E.
    *B. Sample Traffic
        Reports
  ***c. Sample Data Summary
        Package
  ***D. Sample Data Package
  ***E. Complete SDG File
 ****
*****F. Quality
        Assurance
        Plan
3 days after
receipt of last
sample in Sample
Delivery Group
(SDG).**

14 days after
receipt of last
sample in SDG.

14 days after
receipt of last
sample in SDG.

14 days after
receipt of last
sample in SOG.

60 days after
contract award,
and as required
in Exhibit E.
                    X .
                    As directed
  ***G. Data in Computer-
        Readable Form
14 days after
receipt of last
sample in SDG.
                                   B-3B
                           OLM01.3  2/91

-------
               Item
                      No.
                    Copies
Delivery
Schedule
   Distribution
(1)   (2)   (3)
     H. GC/MS Tapes
     I. Extracts
                    Lot     Retain  for  365 days
                            after data  submis-
                            sion, or submit with-
                            in  7 days after
                            receipt of  written
                            request by  APO and/or
                            EMSL/LV.

                    Lot     Retain  for  365 days
                            after data  submis-
                            sion, or submit with-
                            in  7 days after
                            receipt of  written
                            request by  APO or SMO.
    As Directed
    As Directed
Distribution:
(1)  Sample Management Office (SMO)
(2)  Region-Client (Technical Project Officer)
(3)  EMSL-LV
(4)  NEIC
**
***
****
*****
Also required in the Sample Data Summary Package.

Sample Delivery Group (SDG) is a group of samples within a Case, received
over a period of 14 days or less (7 days or less for 14-day data turnaound
contracts) and not exceeding 20 samples.  Data for all samples in the SDG
are due concurrently.  The date of delivery of the SDG or any samples
within the SDG is the date that all samples have been delivered.  (See
Exhibit A for further description).

Concurrent delivery required.  Delivery shall be made such that all
designated recipients receive the item on the same calendar day.

Complete SDG File will contain the original sample data package plus all
of the original documents described under Complete SDG File paragraph E.

See Exhibit E for a more detailed description.
NOTE:  As specified in the Contract Schedule (Section G, Government Furnished
Supplies and Materials), unless otherwise instructed by the CLP Sample Management
Office, the Contractor shall dispose of unused sample volume and used sample
bottles/containers no earlier than sixty (60) days following submission of  the
reconciled complete SDG file.
                                    B-4
                                                                OLM01.3  2/91

-------
Distribution Addresses:

(1)  USEPA Contract Lab Program
     Sample Management Office (SMO)
     P.  0. Box 818
     Alexandria,  VA  22313

     For overnight delivery service,  use street address:
     300 North Lee Street
     Alexandria,  VA  22314

(2)  USEPA REGIONS:

     The CLP Sample Management Office,  acting on behalf of the  Administrative
     Project Officer,  will provide the  Contractor with the list of addressees  for
     the ten EPA Regions.   SMO will  provide the Contractor with updated Regional
     address/name lists as necessary throughout the  period of the  contract and
     identify other client recipients on a case-by-case basis.

(3)  USEPA Environmental Monitoring
     Systems Laboratory (EMSL-LV)
     P.  0. Box 93478
     Las Vegas,  NV  89193-3478
     ATTN:  Data Audit Staff

     For overnight delivery service,  use street address:
     944 E.  Harmon, Executive Center
     Las Vegas,  NV  89109
     ATTN:  Data Audit Staff

(4)  USEPA National Enforcement Investigations Center (NEIC)
     Attn: CLP Audit Program
     Denver Federal Center Bldg.  53
     P.O.  Box 25227
     Denver,  CO  80225
                                   B-5                         OLM01.4   3/91

-------
                                  SECTION II

              REPORT DESCRIPTIONS AND ORDER OF DATA DELIVERABLES

The Contractor laboratory shall provide reports and other deliverables as
specified in the Contract Schedule (Performance/Delivery Schedule,  Section
F) .  The required content and form of each deliverable is described in this
Exhibit.

All reports and documentation MUST BE as follows:

     o   Legible,
     o   Clearly  labeled and completed  in accordance with  instructions  in  this
         Exhibit,
     o   Arranged in the order  specified  in this Section,  and
     o   Paginated consecutively  in ascending  order starting from the  SDG
         Narrative.
If submitted documentation does not conform to the above criteria,  the
Contractor will be  required to resubmit such documentation with
deficiency(ies) corrected, at no additional cost to the Agency.

Whenever the Contractor  is required to submit or resubmit data as a result of
an on-site laboratory evaluation, or through a Administrative Project
Officer/Technical Project Officer action, or through a Regional data
reviewer's request,  the  data must be clearly marked as ADDITIONAL DATA and
must be sent to all  three contractual data recipients (SMO, EMSL/LV, and
Region).  A cover letter shall be included which describes what data is being
delivered, to which  EPA  Case(s) it pertains,  and who requested the data.

Whenever the Contractor  is required to submit or resubmit data as a result of
Contract Compliance  Screening  (CCS) review by SMO, the data must be sent  to
all three contractual data recipients (SMO, EMSL/LV and Region),  and in all
three instances must be  accompanied by a color-coded COVER SHEET (Laboratory
Response To Results  of Contract Compliance Screening) provided by SMO.

Section III of this  Exhibit contains forms instructions to assist the
Contractor in accurately providing the Agency with all required data.
Section IV contains  copies of  the required data reporting forms in
Agency-specified formats.  Data elements with field parameters for reporting
data in computer readable form are contained  in Exhibit H.

Descriptions of  the  requirements for each deliverable item cited in the
Contract Performance/Delivery  Schedule (Contract Schedule, Section F)  are
specified in A-H of this Section.  Items submitted concurrently MUST BE
arranged in the order listed.  Additionally,   the components of each item  MUST
BE arranged in the  order presented in this Section when the item is
submitted.

A.   Quality Assurance Plan and Standard Operating Procedures

     See Exhibits E and  F for  requirements.
                                    B-6                         OLM01.3  2/91

-------
B.   Sample Traffic Reports

     Original Sample Traffic Report page marked "Lab  Copy  for  Return to  SMO"
     with lab receipt information and signed in original Contractor
     signature,  for each sample in the Sample Delivery Group.

     Traffic Reports (TRs)  shall be submitted in Sample Delivery Group (SDG)
     sets (i.e.,  TRs for all samples in an SDG shall  be clipped together),
     with an SDG Cover Sheet attached.

     The SDG Cover Sheet shall contain the following  items:

     o   Lab  name.

     o   Contract  number.
     o   Sample Analysis  Price  -  full sample  price from contract.

     o   Case Number.

     o   List of EPA sample  numbers  of  all  samples in  the SDG,  identifying  the
         first and last  samples  received,  and their dates of receipt  (LRDs).
         NOTE: When more  than  one  sample  is  received  in the first or  last SDG
         shipment,  the "first"  sample received would be the lowest sample
         number  (considering both  alpha and  numeric designations); the "last"
         sample received  would be  the highest sample number (considering both
         alpha and numeric  designations).

     In addition,  each Traffic Report must be clearly marked  with the SDG
     Number,  the sample  number of the first sample  in the  SDG (as described
     in the following paragraph).   This information should be entered below
     the Lab Receipt: Date on the TR.  In addition,  the TR  for the last sample
     received in the £DG must be clearly marked "SDG  - FINAL  SAMPLE."

     The EPA sample number of the first sample received in the SDG is the' SDG
     number.   When several samples are received together  in the first SDG
     shipment,  the SDG number shall be the lowest sample number (considering
     both alpha  and numeric designations) in the first group  of samples
     received under the SDG.   The SDG number is also reported on all data
     reporting forms (see Section III,  Forms Instruction Guide).

     If samples  are received at the laboratory with multi-sample Traffic
     Reports (TRs), all  the samples on one multi-sample TR may not
     necessarily be in the same SDG.  In this instance, the laboratory must
     make the appropriate number of photocopies of the TR,  and submit one
     copy with each SDG  cover sheet.

C.   Sample Data Summary Package

     As specified in the Delivery Schedule, one Sample Data Summary Package
     shall be delivered to SMO concurrently with delivery  of  other required
     sample data.   The Sample Data Summary Package  consists of copies of
     specified items from the Sample Data Package.  These  items are listed
     below and are described under part D, Sample Data Package.
                                    B-7                                OLM01.0

-------
     The Sample Data Summary Package  shall  be  ordered as  follows  and  shall  be
     submitted separately (i.e.,  separated  by  rubber bands,  clips or  other
     means)  directly preceding the  Sample Data Package.   Sample data  forms
     shall be arranged in increasing  EPA sample number order,  considering
     both letters and numbers.   For example, BE400  is a lower  sample  number
     than BF100,  as E precedes F in the  alphabet.

     The Sample Data Summary Package  shall  contain  data for  samples  in one
     Sample  Delivery Group of the Case,  as  follows:

     1.    SDG Narrative

     2.    By fraction (VOA,  SV,  PEST) and by sample within each fraction -
          tabulated target compound results (Form I) and  tentatively
          identified compounds  (Form  I,  TIC) (VOA and SV  only)

     3.    By fraction (VOA,  SV,  PEST) -  surrogate spike analysis  results (Form
          II) by  matrix  (water and/or soil) and for soil, by concentration
          (low or medium)

     4.    By fraction (VOA,  SV,  PEST) -  matrix spike/matrix  spike duplicate
          results (Form  III)

     5.    By fraction (VOA,  SV,  PEST) -  blank  data  (Form  IV) and  tabulated
          results (Form  I)  including  tentatively identified  compounds (Form I,
          TIC)  (VOA and  SV only)

     6.    By fraction (VOA,  SV  only)  - internal standard  area  data (Form VIII)

D.    Sample  Data  Package

     The Sample Data Package is divided  into the five major  units described
     below.  The last three units are  each specific  to an  analytical  fraction
     (volatiles,  semivolatiles,  pesticides/Aroclors).  If the  analysis of  a
     fraction is  not required,  then that fraction-specific unit  is not
     required as  a deliverable.

     The Sample Data Package shall  include  data for analyses of  all  samples
     in  one  Sample Delivery Group,  including field  samples,  reanalyses,
     blanks, matrix spikes, and matrix spike duplicates.

     The Contractor shall retain a  copy  of  the sample Data Package for 365
     days after final acceptance of data.   After this time,  the  Contractor
     may dispose  of the  package.

     1.    SDG Narrative

          This document  shall be  clearly labeled "SDG Narrative"  and shall
          contain:   laboratory name;  Case number; sample  numbers  in the Sample
          Delivery Group (SDG),  differentiating between initial  analyses and
          re-analyses; SDG number;  Contract number; and detailed  documentation
          of any  quality control, sample, shipment  and/or analytical  problems
          encountered in processing the  samples reported  in  the  data  package.
                                   B-8                                OLM01.0

-------
     Whenever data from sample re-analyses are submitted,  the Contractor
     shall state in the SDG Narrative for each re-analysis,  whether it
     considers the re-analysis to be billable, and if so,  why.

     The Contractor must also include any problems encountered:  both
     technical and administrative, the corrective actions  taken, and
     resolution and an explanation for all flagged edits (i.e.,  manual
     edits) on quantitation lists.

     The Contractor must also list the pH determined for- each water
     sample submitted for volatiles analysis.  This information may
     appear as a simple list or table in the SDG Narrative.   The purpose
     of this pH determination is to ensure that all volatiles samples
     were acidified in the field.  No pH adjustment is to  be performed by
     the Contractor on water samples for volatiles analysis.

     The SDG Narrative shall contain the following statement, verbatim:
     "I certify that this data package is in compliance with the terms
     and conditions of the contract, both technically and for
     completeness, for other than the conditions detailed above.  Release
     of the data contained in this hardcopy data package and in the
     computer-readable data submitted on diskette has been authorized by
     the Laboratory Manager or his designee, as verified by the following
     signature."  This statement shall be directly followed by signature
     of the Laboratory Manager or his designee with a typed line below it
     containing the signer's name and title, and the date  of signature.

     Additionally, the SDG Narrative itself must be signed in original
     signature by the Laboratory Manager or his designee and dated.

2.   Traffic Reports

     A copy of the Sample Traffic Reports submitted in Item A for all of
     the samples in the SDG.  The Traffic Reports shall be arranged in
     increasing EPA sample number order, considering both letters and
     numbering in ordering samples.  Copies of the SDG cover sheet is to
     be included with the copies of the Traffic Reports.

     If samples are received at the laboratory with multi-sample Traffic
     Reports (TRs), all the samples on one multi-sample TR may not
     necessarily be in the same SDG.  In this instance, the laboratory
     must make the appropriate number of photocopies of the TR so that a
     copy is submitted with each data package to which it applies.  In
     addition,  in any instance where samples from more than one multi-
     sample TR are in the same data package, the laboratory must submit a
     copy of the SDG cover sheet with copies of the TRs.

3.   Volatiles Data

     a.    QC  Summary

          (1)  System Monitoring  Compound Summary (Form II VOA)

          (2)  Matrix Spike/Matrix Spike  Duplicate  Summary  (Form III VOA)

                               B-9                         OLM01.2  1/91

-------
(3)  Method Blank Summary (Form IV VOA)

     (If more than a single form is necessary,  forms must be
     arranged in chronological order by  date of analysis of the
     blank, by instrument.)

(4)  GC/MS Instrument Performance Check  (Form V VOA)

     In chronological order;  by instrument.

(5)  Internal Standard Area and RT Summary (Form VIII VOA)

     In chronological order;  by instrument.

Sample Data

Sample data shall be arranged in packets with the Organic
Analysis Data Sheet (Form I VOA, including Form I VOA-TIC),
followed by the raw data for volatile samples.   These sample
packets should then be placed in increasing EPA sample number
order, considering both letters and numbers in ordering
samples.

(1)  Target Compound Results  - Organic Analysis Data Sheet
     (Form I VOA).

     Tabulated results (identification and quantitation) of the
     specified target compounds (Exhibit C).  The validation
     and release of these results is authorized by a specific,
     signed statement in the  SDG Narrative (reference D.I).   In
     the event that the Laboratory Manager cannot validate all
     data reported for each sample,  the  Laboratory Manager
     shall provide a detailed description of the problems
     associated with the sample in the SDG Narrative. '

(2)  Tentatively Identified Compounds (Form I VOA-TIC).

     This form must be included even if  no compounds are found.
     If so, indicate this on the form by entering "0" in the
     field for "Number found."

     Form I VOA-TIC is the tabulated list of the highest
     probable match for up to 10 organic compounds that are not
     system monitoring compounds and are not listed in Exhibit
     C (TCL).   It includes the CAS (Chemical Abstracts Service)
     Registry Number,  tentative identification, and estimated
     concentration.

(3)  Reconstructed total ion chromatograms (RIG) for each
     sample or sample extract.

     RICs must be normalized to the largest nonsolvent
     component,  and must contain the following  header
     information:

                     B-10                        OLM01.2  1/91

-------
     o  EPA sample number.
     o  Date and time of analysis.

     o  GC/MS instrument ID.

     o  Lab file ID.
     Internal  standards  and system monitoring compounds  are  to
     be labeled  with  the names  of  compounds,  either  directly
     out from  the peak,  or  on a print-out  of  retention times if
     retention times  are printed over  the  peak.

     If automated data system procedures are  used for
     preliminary identification and/or quantification of the
     target compounds, the  complete data system  report must  be
     included  in all  sample data packages,  in addition to the
     reconstructed  ion chromatogram.   The  complete data  system
     report shall include all of the information listed  below.
     For laboratories which do  not use the automated data
     system procedures,  a laboratory "raw  data sheet,"
     containing  the following information,  must  be included  in
     the sample  data  package in addition to the  chromatogram.

     o  EPA sample number.
     o  Date and time of analysis.

     o  RT or scan number of identified target compounds.

     o  Ion used for quantitation with measured area.

     o  Copy of area table from data system.

     o  GC/MS instrument ID.
     o  -Lab file ID.

     In all instances where the data system report has been
     edited, or  where manual integration or quantitation has
     been  performed,  the GC/MS  operator must  identify such
     edits or  manual  procedures by initialing and dating the
     changes made to  the report.

(4)   For each  sample, by each compound identified, the
     following shall  be  included in the data  package:

     (a)   Copies of raw  spectra and copies of
          background-subtracted mass spectra  of  target
          compounds listed  in Exhibit  C (TCL)  that are
          identified  in  the  sample  and  corresponding
          background-subtracted target  compound  standard mass
          spectra.   Spectra must be  labeled with  EPA sample
          number, lab file  ID,   date  and time  of  analysis,  and
          GC/MS  instrument  ID;   compound names  must be  clearly
          marked on all  spectra.
                    B-ll                               OLM01.0

-------
     (b)  Copies of mass spectra of  organic  compounds  not
          listed in Exhibit C (TCL),  Tentatively Identified
          Compounds,  with associated best-match spectra (three
          best matches), labeled as  in (4)(a)  above.

Standards Data

(1)  Initial Calibration Data (Form VI VGA)  -  in order by
     instrument, if more than one instrument used.

     (a)  VOA standard(s) reconstructed ion  chromatograms  and
          quantitation reports (or legible facsimile)  for  the
          initial (five point) calibration,  labeled as in  b.(3)
          above.  Spectra are not required.

     (b)  All initial calibration data that  pertain to samples
          in the data package must be included,  regardless of
          when it was performed and for which  Case.  When  more
          than one initial calibration is performed,  the data
          must be put in chronological order,  by instrument.

(2)  Continuing Calibration (Form VII VOA)  - in order by
     instrument, if more than one instrument used.

     (a)  VOA standard(s) reconstructed ion  chromatograms  and
          quantitation reports (or legible facsimile)  for  all
          continuing (12 hour) calibrations, labeled as in
          b.(3)  above.   Spectra are not required.

     (b)  When more than one continuing calibration is
          performed,  forms must be in chronological order,
          within fraction and instrument.

Raw QC Data

(1)  BFB  (for each 12-hour period,  for each GC/MS  system
     utilized)

     (a)  Bar graph spectrum,  labeled as in  b.(3) above.

     (b)  Mass listing,  labeled as in b.(3)  above.

     (c)  Reconstructed total ion chromatogram (RIC),  labeled
          as in b.(3)  above.

(2)  Blank Data - in chronological order.  NOTE:  This order  is
     different from that used for samples.

     (a)  Tabulated results (Form I  VOA).

     (b)  Tentatively Identified Compounds (Form I  VOA-TIC)
          even if none  found.
                     B-12                               OLM01.0

-------
               (c)  Reconstructed ion chroraatogram(s) and quantitation
                   report(s) or legible facsimile (GC/MS), labeled as in
                   b. (3) above.

               (d)  Target compound spectra with lab generated standard,
                   labeled as in b.(4) above.  Data systems which are
                   incapable of dual display shall provide spectra in
                   the following order:

                   o   Raw target compound spectra.

                   o   Enhanced or background subtracted spectra.

                   o   Laboratory generated standard spectra.

               (e)  GC/MS library search spectra for Tentatively
                   Identified Compounds (TIC), labeled as in b.(4)
                   above.

               (f)  Quantitation/Calculation of Tentatively Identified
                   Compound(s) (TIC) concentrations.

          (3)   Matrix Spike Data

               (a)  Tabulated results (Form I VOA) of target compounds.
                   Form I VOA-TIC not required.

               (b)  Reconstructed ion chromatogram(s) and quantitation
                   report(s) or legible facsimile (GC/MS), labeled as in
                   b.(4) above.  Spectra not required.

          (4)   Matrix Spike Duplicate Data

               (a)  Tabulated results (Form I VOA) of target compounds.
                   Form I VOA-TIC not required.

               (b)  Reconstructed ion chromatogram(s) and quantitation
                   report(s) or legible facsimile (GC/MS), labeled as in
                   b.(4) above.  Spectra not required.

4.   Semivolatiles Data

     a.    QC Summary

          (1)   Surrogate Percent Recovery Summary  (Form  II SV)

          (2)   Matrix Spike/Matrix Spike Duplicate Summary (Form  III  SV)

          (3)   Method Blank Summary  (Form IV  SV)

               (If more than a single form  is necessary,  forms must be
               arranged in chronological order by  date of analysis of Che
               blank.)
                              B-13                               OLM01.0

-------
     (4)   GC/MS  Instrument  Performance Check  (Form V SV)

          In  chronological  order; by  instrument.

     (5)   Internal  Standard Area  and  RT Summary  (Form VIII SV)

          In  chronological  order; by  instrument.

b.   Sample Data

     Sample data shall be arranged  in packets with  the  Organic
     Analysis Data Sheet (Form I  SV,  including Form I SV-TIC),
     followed by the raw data for semivolatile samples.   These
     sample packets must then be  placed  in increasing EPA sample
     number order,  considering both letters and  numbers in ordering
     samples.

     (1)   Target Compound Results - Organic Analysis Data Sheet
          (Form  I SV-1,  SV-2).

          Tabulated results (identification and  quantitation)  of the
          specified target  compounds  (Exhibit C).   The  validation
          and release  of these  results is  authorized by a specific,
          signed statement  in the SDG Narrative  (reference D.I).   In
          the event that the  Laboratory Manager  cannot  validate  all
          data reported for each  sample, the  Laboratory Manager
          shall  provide a detailed  description of the problems
          associated with the sample  in the SDG  Narrative.

     (2)   Tentatively  Identified  Compounds  (Form I  SV-TIC).

          This form must be included  even  if  no  compounds are  found.
          If  so, indicate this  on the form by entering  "0"  in  the
          field  for "Number found".

          Form I SV-TIC is  the  tabulated list of the highest
          probable  match for  up to  20 of the  non-surrogate  organic
          compounds not listed in Exhibit  C (TCL).   It  includes  the
          CAS (Chemical Abstracts Service)  Registry Number,
          tentative identification, and estimated concentration.

     (3)   Reconstructed total ion chromatograms  (RIG) for each
          sample, sample extract, standard, blank,  and  spiked
          sample.

          RICs must be normalized to  the largest nonsolvent
          component, and must contain the  following header
          information:

          o   EPA sample  number.

          o   Date  and  time of  analysis.
                         B-14                         OLM01.2  1/91

-------
     o  GC/MS instrument ID.

     o  Lab file ID.
     Internal  standards  and  surrogates compounds are  to be
     labeled with  the  names  of  compounds, either directly out
     from  the  peak,  or on a  print-out of retention  times if
     retention times are printed  over the peak.

     If automated  data system procedures are used for
     preliminary identification and/or quantification of the
     target compounds, the complete data system report must be
     included  in all sample  data  packages,  in  addition to the
     reconstructed ion chromatogram.  The complete  data system
     report shall  include all of  the information listed below.
     For laboratories  which  do  not use the  automated  data
     system procedures,  a laboratory "raw data sheet,"
     containing  the  following information,  must be  included in
     the sample  data package in addition to the chromatogram.

     o  EPA sample number.

     o  Date and time of analysis.

     o  RT or scan number of identified target compounds.

     o  Ion used for quantitation with measured area.

     o  Copy of area table  from data system.
     o  GC/MS instrument ID.
     o  Lab file ID.

     In all instances  where  the data system report  has been
     edited, or  where  manual integration or quantitation has
     been  performed, the GC/MS  operator must identify such
     edits or  manual procedures by initialing  and dating the
     changes made  to the report.

(4)   For each  sample,  by each compound identified,  the
     following shall be  included  in the data package:

     (a)   Copies of  raw  spectra and copies  of
          background-subtracted mass spectra of target
          compounds  listed in Exhibit C (TCL)  that  are
          identified in  the  sample and corresponding
          background-subtracted target compound standard mass
          spectra.   Spectra  must be labeled with EPA  sample
          number, lab  file ID,   date and time of analysis, and
          GC/MS  instrument ID;   compound names  must  be  clearly
          marked on all  spectra.

     (b)   Copies of mass spectra of nonsurrogate organic
          compounds not  listed  in Exhibit C (TCL),  Tentatively
          Identified Compounds  with associated best-match
          spectra (three best matches), labeled as  in (4)(a)
          above.


                    B-15                               OLM01.0

-------
c.   Standards Data

     (1)  Initial Calibration Data (Form VI  SV-1,  SV-2)  -  in order
          by instrument,  if more than one instrument  used.

          (a)   Semivolatile standard(s)  reconstructed ion
               chromatograms and quantitation reports (or  legible
               facsimile)  for the initial (five point) calibration,
               labeled  as  in b.(3)  above.  Spectra are not  required.

          (b)   All  initial calibration data  that pertain to samples
               in the data package must  be included,  regardless of
               when it  was performed  and for which Case.  When more
               than one initial  calibration  is performed,  the data
               must be  put in chronological  order, by instrument.

     (2)  Continuing Calibration (Form VII SV-1, SV-2)  - in order  by
          instrument,  if  more than one instrument  used.

          (a)   Semivolatile standard(s)  reconstructed ion
               chromatograms and quantitation reports (or  legible
               facsimile)  for all continuing (12 hour) calibrations,
               labeled  as  in b.(3)  above.  Spectra are not  required.

          (b)   When more  than one continuing calibration is
               performed,  forms  must  be  in chronological order, by
               instrument.

     (3)  Semivolatile  GPC Calibration Data  - UV detector  traces
          showing peaks that correspond  to the compounds in the
          Semivolatile  GPC calibration mixture.  Traces  must be
          labeled with  GPC column identifier, date of calibration,
          and  with compound names labeled either directly  out from
          the  peak,  or  on a printout  of  retention  times, if
          retention times  are printed over the peak.   Do not include
          Form IX Pest-2,  as the compounds used on that  form are not
          appropriate  for Semivolatile sample extracts.

d.   Raw QC Data

     (1)  DFTPP (for each 12-hour period,  for each GC/MS system
          utilized)

          (a)   Bar  graph spectrum,  labeled as in b(3) above.

          (b)   Mass  listing,  labeled  as  in b.(3) above.

          (c)   Reconstructed total  ion chromatogram (RIC),  labeled
               as in b.(3)  above.
                         B-16                               OLM01.0

-------
          (2)  Blank Data -  in chronological  order by  extraction  date.
               NOTE:   This order  is  different from that used  for  samples.

               (a)  Tabulated  results  (Form I  SV-1, SV-2)

               (b)  Tentatively Identified Compounds (Form I  SV-TIC)  -
                   even  if  none  found.

               (c)  Reconstructed ion  chromatogram(s)  and quantitation
                   report(s)  or  legible facsimile (GC/MS), labeled as  in
                   b(3)  above.

               (d)  Target compound  spectra with lab generated standard,
                   labeled  as in b(4) above.  Data systems which are
                   incapable  of  dual  display shall provide spectra in
                   the following order:

                   o   Raw target compound spectra.

                   o   Enhanced or background subtracted spectra.

                   o   Laboratory generated standard spectra.

               (e)  GC/MS library search spectra for Tentatively
                   Identified Compounds (TIC), labeled as in b(4) above.

               (f)  Quantitation/Calculation  of Tentatively Identified
                   Compound(s) (TIC)  concentrations

          (3)  Matrix  Spike  Data

               (a)  Tabulated  results  (Form I) of target compounds. Form
                   1  SV-TIC not  required.

               (b)  Reconstructed ion  chromatogram(s)  and quantitation
                   report(s)  or  legible facsimile (GC/MS), labeled as  in
                   b(3)  above.   Spectra not  required.

          (4)  Matrix  Spike  Duplicate  Data

               (a)  Tabulated  results  (Form I  SV-1, SV-2) of  target
                   compounds.  Form 1 SV-TIC  not required.

               (b)  Reconstructed ion  chromatogram(s)  and quantitation
                   report(s)  or  legible facsimile (GC/MS), labeled as  in
                   b(3) above.   Spectra not  required.

5.   Pesticide/Aroclor Data

     a.    QC  Summary

          (1)   Surrogate  Percent  Recovery Summary  (Form II PEST)

          (2)   Matrix  Spike/Matrix Spike Duplicate Summary (Form  III
               PEST)
                              B-17                         OLM01.2  1/91

-------
     (3)   Method Blank Summary  (Form  IV  PEST)

          (If more  than a  single  form is necessary,  forms must  be
          arranged  in chronological order  by date  of analysis of  the
          blank.)

b.   Sample Data

     Sample data shall be  arranged in packets  with the Organic
     Analysis Data Sheet (Form  I  PEST),  followed by the raw data  for
     pesticide samples. These  sample packets  should then be placed
     in increasing EPA sample number  order,  considering both letters
     and numbers in ordering samples.

     (1)   Target Compound  Results - Organic Analysis Data Sheet
          (Form I PEST).

          Tabulated results  (identification and quantitation) of  the
          specified target compounds  (Exhibit  C).   The validation
          and release of these  results is  authorized by a specific,
          signed statement in the SDG Narrative (reference  D.I).   In
          the event that the Laboratory  Manager cannot validate all
          data reported for  each  sample, the Laboratory Manager
          shall provide a  detailed description of  the problems
          associated with  the sample  in  the SDG Narrative.

     (2)   Copies of pesticide chromatograms.

          All chromatograms  must  be labeled with the following
          information:

          o   EPA Sample Number.

          o   Volume  injected (uL).

          o   Date and  time of injection.

          o   GC column identification (by stationary phase and
              internal diameter).

          o   GC instrument identification.

          o   Positively identified compounds must be  labeled with
              the names of compounds,  either directly  out from  the
              peak, or on a print-out of retention times if
              retention times are printed over  the peak.

     (3)   Copies of pesticide chromatograms from second GC  column,
          labeled as in (2)  above.

     (4)   GC Integration report or data  system printout.

     (5)   Manual work sheets.
                         B-18                         OLM01.2  1/91

-------
     (6)   If  pesticide/Aroclors are confirmed by GC/MS,  the
          Contractor  shall submit copies of reconstructed ion
          chromatograms, raw spectra and background-subtracted mass
          spectra  of  target compounds listed in Exhibit C (TCL) that
          are identified in the sample and corresponding
          background-subtracted TCL standard mass spectra.  Compound
          names must  be clearly marked on all spectra.  For
          multicoraponent pesticides/Aroclors confirmed by GC/MS, the
          Contractor  shall submit mass spectra of 3 major peaks of
          multicomponent compounds from samples and standards.

c.   Standards Data

     (1)   Initial  Calibration of Single Component Analytes (Form VI
          PEST-1 and  PEST-2) - all GC columns, all instruments, in
          chronological order by GC column and instrument.

     (2)   Initial  Calibration of Multicomponent Analytes  (Form VI
          PEST-3)  - all GC columns, all instruments,  in
          chronological order by GC column and instrument.

     (3)   Analyte  Resolution Summary (Form VI PEST-4) - all GC
          columns  and instruments, in chronological order by GC
          column and  instrument.

     (4)   Calibration Verification Summary (Form VII  PEST-1) - for
          all Performance Evaluation Mixtures and Instrument blanks,
          on  all GC columns and instruments, in chronological order
          by  GC column and instrument.

     (5)   Calibration Verification Summary (Form VII  PEST-2) - for
          all mid  point concentrations of Individual  Standard
          Mixtures A  and B and Instrument blanks used for
          calibration verification, on all GC columns and
          instruments, in chronological order by GC column and
          instrument.

     (6)   Analytical  Sequence (Form VIII PEST) - all  GC columns and
          instruments, in chronological order by GC column and
          instrument.

     (7)   Florisil Cartridge Check (Form IX PEST-1) - for all lots
          of  cartridges used to process samples in the SDG.

     (8)   Pesticide GPC Calibration (Form IX PEST-2)  - all GPC
          columns,  in chronological order by calibration  date.
      i
     (9)   Pesticide Identification Summary for Single Component
          Analytes (Form X PEST-1) - for all samples  with positively
          identified  single component analytes, in order  by
          increasing  EPA sample number.
                         B-19                         OLM01.2  1/91

-------
(10) Pesticide Identification Summary for Multicomponent
     Analytes (Form X PEST-2)  -  for  all  samples with  positively
     identified multicomponent analytes,  in  order  by  increasing
     EPA sample number.

(11) Chromatograms and data system printouts are  required for
     all standards including the following:

     o  Resolution Check Mixture.

     o  Performance Evaluation Mixtures, all.

     o  Individual Standard Mixture A, at three
        concentrations, each initial calibration.

     o  Individual Standard Mixture B, at three
        concentrations, each initial calibration.
     o  All multicomponent analytes (Toxaphene and Aroclors),
        each  initial calibration.
     o  All mid point concentrations of Individual Standard
        Mixtures  A and B used for calibration verification.

     o  Florisil  cartridge check solution, all lots.

     o  Pesticide GPC Calibration Check Solution, all
        calibrations relating to samples in the SDG.

     o  All multicomponent analyte standards analyzed for
        confirmation.

(12) A printout of retention times and corresponding peak areas
     or peak heights  must  accompany  each chromatogram.   In
     addition,  all chromatograms are required to  be  labeled
     with  the following:

     o  EPA Sample Number for the standard, i.e., INDAl,
        INDA2, etc.  (See Forms Instructions for  details.)

     o  Label all standard peaks for all individual  compounds
        either directly out from the peak or on the  printout
        of retention times  if retention times are printed  over
        the peak.

     o  Total nanograms injected for each standard.

     o  Date and  time of injection.

     o  GC column identification (by stationary phase and
        internal  diameter).

     o  GC instrument identification.

(13) Pesticide GPC Calibration Data - UV detector traces
     showing peaks that  correspond to the compounds  in the
     pesticide  GPC calibration mixture.   Traces must  be  labeled
     with  GPC column  identifier,  date of calibration, and with
     compound names  labeled either directly  out  from the peak,
     or on a printout of retention times,  if retention times
     are printed over the  peak.
                    B-20                         OLM01.2  1/91

-------
          d.   Raw QC Data

              (1)  Blank Data - in chronological order, by type of blank
                   (method, instrument, sulfur clean up).  NOTE:  This order
                   is different from that used for samples.

                   (a)  Tabulated results (Form I PEST).

                   (b)  Chromatogram(s) and data system printout(s)  (GC)  for
                        each GC column and instrument used for analysis,
                        labeled as in b.(2) above.

              (2)  Matrix Spike Data

                   (a)  Tabulated results (Form I PEST) of target compounds.

                   (b)  Chromatogram(s) and data system printout(s)  (GC),
                        labeled as in b.(2) above.

              (3)  Matrix Spike Duplicate Data

                   (a)  Tabulated results (Form I PEST) of target compounds.

                   (b)  Chromatogram(s) and data system printout(s)  (GC),
                        labeled as in b.(2) above.

E.    Complete SPG File

     As  specified in  the Delivery Schedule,  one Complete  SDG  File (CSF)
     including the  original Sample Data Package shall be  delivered to  the
     Region concurrently with  delivery of the  Sample  Data Package to SMO and
     EMSL-LV.  The  contents of the CSF will be numbered according to the
     specifications described  in Section III and IV of Exhibit B.  The
     Document Inventory Sheet,  Form DC-2,  is contained in Section IV.  The
     CSF will contain all original documents where  possible.   No copies will
     be  placed in the CSF unless the originals are  bound  in a logbook  which
     is  maintained  by the laboratory.   The CSF will contain all original
     documents specified in Section III and IV,  and Form  DC-2 of Exhibit B of
     the SOW.

     The CSF will consist of the following original documents in addition  to
     the documents  in the Sample Data Package:

     1.    Original Sample Data Package

     2.    A completed and signed Document  Inventory Sheet (Form DC-2).
                                                                      i
     3.    All  original shipping  documents,  including,  but not limited  to,  the
          following documents:

          a.  EPA Chain of Custody Record.

          b.  Airbills.
                                   B-21                               OLM01.0

-------
          c.  EPA  Traffic Reports.
          d.  Sample Tags (if present) sealed in plastic bags.

     4.   All original receiving documents,  including,  but not limited to,  the
          following documents:

          a.  Form DC-1.
          b.  Other receiving forms or copies of receiving logbooks.
          c.  SDG  Cover Sheet.

     5.   All original laboratory records,  not  already  submitted in the Sample
          Data Package,  of  sample transfer,  preparation and analysis,
          including,  but not limited to,  the following  documents:

          a.  Original preparation and analysis forms or copies of preparation
              and  analysis  logbook pages.
          b.  Internal sample and sample extract transfer chain-of-custody
              records.
          c.  Screening records.
          d.  All  instrument output,  including strip charts  from screening
              activities.
     6.   All other original SDG-specific  documents in  the possession  of the
          laboratory,  including,  but not limited  to,  the following  documents:

          a.  Telephone contact logs.

          b.  Copies of personal  logbook pages.

          c.  All  hand written case-specific notes.

          d.  Any  other case-specific documents not covered by  the above.

     NOTE:  All Case-related documentation may be used  or admitted as
     evidence in subsequent legal proceedings.  Any other Case-specific
     documents generated after the CSF is  sent to EPA,  as well as copies that
     are altered in any fashion,  are also  deliverables  to EPA.  (Original to
     the Region, and copies to SMO and EMSL-LV).

     If the laboratory does submit SDG-specific documents to EPA after
     submission of the CSF, the documents  should  be numbered as an addendum
     to the CSF and a revised DC-2 form should be submitted, or the documents
     should be numbered as  a new CSF and a new DC-2 form should be submitted
     to the Regions only.

F.   Data in Computer-Readable Form

     The Contractor shall provide a computer-readable copy of the data on
     data reporting Forms I-X for all samples in  the Sample Delivery Group,
     as specified in the Contract Performance/Delivery  Schedule.
     Computer-readable data deliverables shall  be submitted on IBM or
     IBM-compatible,  5.25 inch floppy double-sided, double density 360
     K-byte, a high density 1.2 M-byte or  a 3.5 inch double-sided double
     density 720 K-byte or  1.44 M-byte diskette.


                                   B-22                               OLM01.0

-------
     When submitted,  diskettes shall be packaged and shipped in such a manner
     that the diskette(s) cannot be bent or folded,  and will not be  exposed
     to extreme heat or cold or any type of electromagnetic  radiation.   The
     diskette(s) must be included in the same shipment as  the hardcopy data
     and shall, at a minimum,  be enclosed in a diskette mailer.   The data
     shall be recorded in ASCII, text file format,  and shall adhere  to the
     file, record and field specifications listed in Exhibit H,  Data
     Dictionary and Format for Data Deliverables in Computer-Readable Format.

G.   GC/MS Tapes

     See Exhibit E for requirements.

H.   Extracts

     The Contractor shall preserve sample extracts  at 4°C  (+2"C) in
     bottles/vials with Teflon-lined septa.   Extract bottles/vials shall be
     labeled with EPA sample number,  Case number and Sample  Delivery Group
     (SDG) number.   A logbook of stored extracts shall be  maintained,  listing
     EPA sample numbers and associated Case and SDG numbers.

     The Contractor is required to retain extracts  for 365 days  following
     data submission.   During that time,  the Contractor shall submit extracts
     and associated logbook pages within seven days following receipt of a
     written request from the Administrative Project Officer or the  Sample
     Management Office.
                                   B-23                         OLM01.2   1/91

-------
    SECTION  III






FORMS INSTRUCTIONS
    B-24
                                       OLM01.0

-------
                            SECTION  III


                      FORM INSTRUCTION GUIDE


This section includes specific instructions -for the completion of all
required forms.  Each of the forms is specific to a given fraction
(volatile, semivolatile, pesticide/Aroclor), and in some instances
specific to a given matrix  (water or soil) within each fraction.  The
contractor shall submit only those forms pertaining to the fractions
analyzed for a given sample or samples.  For instance, if a sample is
scheduled for volatile analysis only, provide only VOA forms.  There
are two pages relating to the semivolatile fraction for Forms I, VI,
VII, and VIII.  Whenever semivolatiles are analyzed and one of the
above-named forms is required, both pages (SV-1 and SV-2) must be
submitted.  These instructions are arranged in the following order:

A.  General  Information and Header Information

B.  Organic  Analysis Data Sheets  (Form I,  All  Fractions)

C.  System Monitoring Compound Recovery (Form  II  VOA)

D.  Surrogate  Recovery (Form II,  SV and PEST)

E.  Matrix Spike/Matrix Spike Duplicate Recovery  (Form III,  All
    Fractions)

F.  Method Blank Summary (Form IV,  All Fractions)

G.  GC/MS  Instrument Performance  Check (Form V VOA and SV)

H.  GC/MS  Initial Calibration Data (Form VI  VOA,  SV-1,  SV-2)

I.  GC  Initial Calibration Data (Form VI PEST-1,  PEST-2,  PEST-3,  PEST-4)

J. .GC/MS  Continuing Calibration  Data (Form VII VOA,  SV-1,  SV-2)

K.  GC/EC  Calibration Verification Summary(Form VII PEST)

L.  Internal Standard Area and RT Summary  (Form VIII VOA and SV)

M.  Pesticide  Analytical Sequence (Form VIII PEST)

N.  Pesticide  Cleanup Procedures  (Form IX  PEST-1,  PEST-2)

0.  Pesticide/Aroclor Identification (Form X PEST-1,  PEST-2)

P.  Sample Log-In Sheet (Form DC-1)

Q.  Document Inventory Sheet (Form DC-2)
                             B-25                         OLM01.2  1/91

-------
A.    General Information and Header Information

      The data reporting forms presented in Section IV have been designed in
      conjunction with the computer-readable data format specified in Exhibit
      H, Data Dictionary and Format for Data Deliverables in
      Computer-Readable Format.  The specific length of each variable for
      computer-readable data transmission purposes is given in the Data
      Dictionary (Exhibit H).   Information entered on these forms must not
      exceed the size of the field given on the form, including such
      laboratory-generated items as Lab Name and Lab Sample ID.

      Note that on the hardcopy forms (Section IV),  the space provided for
      entries is greater in some instances than the length prescribed for the
      variable as written to diskette (see Exhibit H).   Greater space is
      provided on the hardcopy forms for the sake of visual clarity.

      Values must be reported on the hardcopy forms according to the
      individual form instructions in this Section.   For example, results for
      concentrations of VGA target compounds must be reported to two
      significant figures if the value is greater than or equal to 10.
      Values can be written to the diskette file in any format that does not
      exceed the field specification as given in the record specifications
      and discussed in "Record Structure", paragraph 5 of Exhibit H.

      All characters which appear on the data reporting forms presented in
      the contract (Exhibit B, Section IV) must be reproduced by the
      Contractor when submitting data, and the format of the forms submitted
      must be identical to that shown in the contract.   No information may be
      added, deleted, or moved from its specified position without prior
      written approval of the EPA Administrative Project Officer.  The names
      of the various fields and compounds (e.g., "Lab Code," "Chloromethane")
      must appear as they do on the forms in the contract,  including the
      options specified in the form (e.g., "Matrix:   (soil/water)" must
      appear, not just "Matrix").  For items appearing on the uncompleted
      forms (Section IV), the use of uppercase and lowercase letters is
      optional.

      Alphabetic entries made onto the forms by the Contractor shall be in
      ALL UPPERCASE letters (e.g., "LOW", not "Low" or "low").  If an entry
      does not fill the entire blank space provided on the form, null
      characters shall be used to remove the remaining underscores that
      comprise the blank line.  See Exhibit H for more detailed instructions.
      However, do not remove the underscores or vertical bar characters that
      delineate "boxes" on the forms.  The only exception would be those
      underscores at the bottom of a "box" that are intended as a data entry
      line (for instance, on Form 2A, line 30, if data must be entered on
      line 30, it will replace the underscores).

      Six pieces of information are common to the header sections of each
      data reporting form.  They are Lab Name, Contract, Lab Code, Case No.,
      SAS No., and SDG No.  Except as noted below for SAS No., this
      information must be entered on every form and must match on every form.
                                   B-26                               OLM01.0

-------
The "Lab Name" shall be the name chosen by the Contractor to identify
the laboratory.  It may not exceed 25 characters.

The "Lab Code" is an alphabetical abbreviation of up to 6 letters,
assigned by the EPA, to identify the laboratory and aid in data
processing.  This lab code shall be assigned by the EPA at the time a
contract is awarded, and shall not be modified by the Contractor,
except at the direction of the EPA.  If a change of name or ownership
occurs at the laboratory, the lab code will remain the same until the
contractor is directed by the EPA to use another lab code assigned by
the EPA.

The "Case No." is the EPA-assigned Case number associated with the
sample, and reported on the Traffic Report.

The "Contract" is the number of the EPA contract under which the
analyses were performed.

The "SDG No." is the Sample Delivery Group number.  The Sample Delivery
Group (SDG) number is the EPA Sample Number of the first sample
received in the SDG.  When several samples are received together in the
first SDG shipment, the SDG number shall be the lowest sample number
(considering both alpha and numeric designations) in the first group of
samples received under the SDG;

The "SAS No." is the EPA-assigned number for analyses performed under
Special Analytical Services.  If samples are to be analyzed under SAS
only and reported on these forms, then enter SAS No. and leave Case No.
blank.  If samples are analyzed according to the "Routine Analytical
Services" (IFB) protocols and have additional "SAS" requirements, list
both Case No. and SAS No. on all forms.  If the analyses have no SAS
requirements, leave "SAS No." blank.  NOTE:  Some samples in an SDG may
have a SAS No. while others do not.

The other information common to most of the forms is the "EPA Sample
No.". This number appears either in the upper righthand corner of the
form, or as the left column of a table summarising data from a number
of samples.  When the "EPA Sample No." is entered into the
triple-spaced box in the upper righthand corner of Form I, Form IV, or
Form X, it should be entered on the middle line of the three lines that
comprise the box.

All samples, matrix spikes, matrix spike duplicates, blanks, and
standards shall be identified with an EPA Sample Number.  For field
samples, matrix spikes and matrix spike duplicates, the EPA Sample
Number is the unique identifying number given in the Traffic Report
that accompanied that sample.
                             B-27                         OLM01.4  3/91

-------
 In order  to  facilitate data assessment, the following sample suffixes
 must be used:

 XXXXX     -   EPA sample number
 XXXXXMS   -   Matrix spike sample
 XXXXXMSD  -   Matrix spike duplicate sample
 XXXXXRE   -   Re-extracted and re-analyzed sample
 XXXXXDL   -   Sample analyzed at a secondary dilution

 Form VIII Pest requires that all samples analyzed in a given analytical
 sequence be  listed, regardless of whether or not they are part of the
 SDG being reported.  Therefore, use "ZZZZZ" as the EPA Sample No. for
 any sample analysis not associated with the SDG being reported.

 For blanks and standards, the following identification scheme must be
 used as the  "EPA Sample No."

         1.   Volatile blanks shall be identified as VBLK##.

         2.   Semivolatile blanks shall be identified as SBLK##.
         3.   Pesticide/Aroclor method blanks shall be identified as
             PBLK##.

         4.   Pesticide/Aroclor instrument blanks shall be identified as
             PIBLK##.

 The "EPA Sample No." must be unique for each blank within an SDG.
 Within a fraction, a laboratory must achieve this by replacing the two-
 character "##" terminator of the identifier with one or two characters
 or numbers,  or a combination of both.  For example, possible
 identifiers  for volatile blanks would be VBLK1, VBLK2,  VBLKAl, VBLKB2,
 VBLK10, VBLKAB, etc.

 Volatile and  semivolatile standards shall be identified as FSTD###,
 where

        F   -     Fraction (V for volatiles; S for semivolatiles).

        STD -     Indicates a standard.

        ###   -    The concentration in ug/L of volatile standards
                  (i.e., 010, 020, 050, 100, and 200) or the amount
                  injected in ng for semivolatile standards
                  (i.e., 020, 050, 080, 120, and 160).

As for the blank identifiers, these designations will have to be
 concatenated with other information to uniquely identify each standard.
                             B-28                               OLM01.0

-------
 For  pesticide/Aroclor standards, the following scheme shall be used to
 enter  "EPA  Sample Number".

       Name                       EPA Sample Number

       Individual Mix A  (low point)       INDAL##
       Individual Mix A  (mid point)       INDAM##
       Individual Mix A  (high point)      INDAH##
       Individual Mix B  (low point)       INDBL##
       Individual Mix B  (mid point)       INDBM##
       Individual Mix B  (high point)      INDBH##
       Resolution Check                   RESC##
       Performance Evaluation Mixture     PEM##
       Toxaphene                          TOXAPH##
       Aroclor 1016                       AR1016##
       Aroclor 1221                       AR1221##
       Aroclor 1232                       AR1232##
       Aroclor 1242                       AR1242##
       Aroclor 1248                       AR1248##
       Aroclor 1254                       AR1254##
       Aroclor 1260                       AR1260##

 The  permitted mixture of Aroclor 1016 and Aroclor 1260 shall be entered
 as AR1660##.

 The  laboratory must create a unique "EPA Sample No." within an SDG by
 replacing the two-character "##" terminator of the identifier with one
 or two characters or numbers, or a combination of both.

 If the standards are injected onto both GC columns on the same
 instrument  simultaneously, the same EPA Sample Number may be used for
 reporting data for the  standards for both columns.  If simultaneous
 injections  are not made, then the same number may not be used.

 Several other pieces of information are common to many of the Data
 Reporting Forms.  These include Matrix, Sample wt/vol, Level, Lab
 Sample ID,  and Lab File ID.

 For  "Matrix", enter "SOIL" for soil/sediment samples, and enter "WATER"
 for water samples.  NOTE:  The matrix must be spelled out.
 Abbreviations such as "S" or "W" shall not be used.

 For  "Sample wt/vol" enter the number of grams (for soil) or milliliters
 (for water) of sample used in the first blank line, and the units,
 either "G"  or "ML", in  the second blank.

 For volatiles and seraivolatiles, for "Level", enter the determination
 of concentration level made from the mandatory screening of soils.
 Enter  as "LOW" or "MED", not "L" or "M".  All water samples are "LOW"
 level  and shall be entered as such.  Note:  There is no differentiation
between low and medium  soil samples for the Pesticide/Aroclor forms,
 and no level is entered on any of the these forms.
                             B-29                         OLM01.2  1/91

-------
      "Lab Sample ID" is an optional laboratory-generated internal
      identifier.  Up to 12 alpha-numeric characters may be reported here.
      If the contractor does not have a Lab Sample ID,  this field may be left
      blank.

      "Lab File ID" is the laboratory-generated name of the GC/MS data system
      file containing information pertaining to a particular analysis.  Up to
      14 alpha-numeric characters may be used here.

      "Instrument ID" is common to many of the forms, particularly those
      containing calibration data.  The identifier used by the laboratory
      must include some indication of the manufacturer and/or model of the
      instrument, and contain additional characters that differentiate
      between all instrument of the same type in the laboratory.

      "GC Column" and "ID (mm)" are common to various other forms.  These two
      fields are to be used to identify the stationary phase of the GC column
      (previously called GC Column ID),  and the internal diameter of the GC
      column in millimeters (mm).  For packed columns,  convert the ID from
      inches to millimeters as necessary, and enter in the "ID" field.

      Forms II, IV, V, VIII, IX, and X contain a field labeled "page _ of _"
      in the bottom lefthand corner.  If the number of entries required on
      any of these forms exceeds the available space, continue entries on
      another copy of the same fraction-specific form,  duplicating all header
      information. If a second page is required, number the pages
      consecutively, as "page 1 of 2" and "page 2 of 2".  If a second page is
      not required, number the page "page 1 of 1."  NOTE:  These forms are
      fraction-specific, and often matrix-specific within fraction.  For
      example, Form II VOA-1 and Form II VGA-2 are for different data.
      Therefore, &o not number the pages of all six versions of Form II as "1
      of 6, 2 of 6, etc."  Number only pages within a fraction-specific and
      matrix-specific form.

      For rounding off numbers to the appropriate level of precision, observe
      the following common rules.  If the figure following those to be
      retained is less than 5, drop it (round down).  If the figure is
      greater than 5, drop it and increase the last digit to be retained by 1
      (round up).  If the figure following the last digit to be retained
      equals 5, round up if the digit to be retained is odd, and round down
      if that digit is even.

B.     Organic Analysis Data Sheet (Form I. All Fractions)

1.     Form I VOA, Form I SV-1, Form I SV-2, Form I PEST

      This form is used for tabulating and reporting sample analysis  results
      for target compounds.  If all fractions are not requested to be
      analyzed, only the pages specifically required must be submitted.   If
      VOA analysis only is requested, Form I VOA and Form I VOA TIC must be
      submitted.  If the pesticide/Aroclor analysis is the only analysis
      requested, only Form I Pest must be submitted for that sample.
                                   B-30                               OLM01.0

-------
Complete Che header information on each page of Form I required,
according to the instructions in Part A and as follows:

For soil samples analyzed for volatiles, for "% moisture not dec.",
enter the nondecanted percent moisture.  This is the only percent
moisture determination made for volatiles, as the entire contents of
the VGA vial are considered as the sample.  For water samples, leave
this field blank.

For soil samples analyzed for semivolatiles and pesticides/Aroclors,
enter the values for the percent moisture determined during the
analysis.  In the field "decanted (Y/N)", enter "Y" if the sample had
standing water above the soil/sediment that was decanted, or "N" if no
water was decanted off the surface of the sample.  Report percent
moisture (decanted or not decanted) to the nearest whole percentage
point (i.e., 5%, not 5.3%).  Leave these fields blank for Form I for
water samples, method blanks, and instrument blanks.

For volatiles, enter the GC column identifier under "GC Column", and
the internal diameters in millimeters (mm), to two decimal places,
under "ID".  For packed columns, convert the ID from inches to
millimeters as necessary, and enter in this field.

For pesticides/Aroclors, enter the method of extraction as "SEPF" for
separatory funnel, "CONT" for continuous liquid-liquid extraction, or
"SONG" for sonication (soils only).
If gel permeation chromatography, "GPC Cleanup", was performed, enter
"Y" for yes.  Otherwise, enter "N" for no, if GPC was not performed.
NOTE:  GPC is required for all soil samples analyzed for semivolatiles
and pesticides/Aroclors, therefore all soil sample forms will contain
"Y" in this field.

For soil samples only, enter pH for semivolatiles and
pesticides/Aroclors, reported to 0.1 pH units.

"Date Received" is the date of sample receipt at the laboratory, as
noted on the Traffic Report (i.e., the VTSR).  It should be entered as
MM/DD/YY.

"Date Extracted" and "Date Analyzed" should be entered in a similar
fashion.  When continuous liquid-liquid extraction procedures are used
for water samples, enter the date on which the procedure was started
for "Date Extracted".  If separatory funnel (pesticides only) or
sonication procedures are used, enter the date on which the procedure
was completed.  For pesticide/Aroclor samples, the date of analysis
should be the date of the first GC analysis performed.  The date of
sample receipt will be compared with the extraction and analysis dates
of each fraction to ensure that contract holding times were not
exceeded.
                             B-31                         OLM01.2  1/91

-------
For volatiles, if a medium soil sample is analyzed, under the "Soil
Extract Volume" enter total volume of the methanol extract in
microliters.  This volume includes any methanol not collected from the
filtration of the extract through glass wool, and is typically 10000
uL, i.e., the 10 mL of methanol use for the extraction.  If a medium
soil sample is analyzed, enter also the volume of the methanol extract
added  to the reagent water in the purge tube and analyzed, under "Soil
Aliquot Volume".  Enter this volume in microliters (uL) .

For semivolatiles, enter the actual volume of the most concentrated
sample extract, in microliters, under "Concentrated Extract Volume".
This volume typically will be 1000 uL (water), or 500 uL (water and
soil) when GPC was performed,--For pesticides/Aroclors, the volume of
the most concentrated extract typically will be 10000 uL (water),  or
5000 uL (water and soil) when GPC is performed.  For
pesticides/Aroclors, the volume of the most concentrated extract is not
the volume taken through the Florisil and sulfur cleanup steps.  If a
dilution of the sample extract is made in a subsequent analysis, this
volume will remain the same, but the dilution factor will change.

For semivolatiles and pesticides/Aroclors, enter the volume of the
sample extract injected into the GC under "Injection Volume".  Report
this volume in microliters to one decimal place, i.e., 1.0 uL.  Note: A
2.0 microliter injection is required for semivolatile analyses.  If
pesticide/Aroclors are analyzed using two GC columns connected to a
single injection port, the "Injection Volume" must be entered as half
Che volume in the syringe, i.e., assume that the extract injected is
evenly divided between the two columns.

If a sample or sample extract has been diluted for analysis, enter the
"Dilution Factor" as a single number, not a fraction, such as "100.0,"
for a 1 to 100 dilution of the sample.  Enter 0.1 for a concentration
of 10 to 1.  If a sample was not diluted, enter "1.0."  Report dilution
factors to one decimal place.

For positively identified target compounds, the Contractor shall report
the concentrations detected as uncorrected for blank contaminants.

For volatile and semivolatile results, report analytical results to one
significant figure if the value is less than 10, and two significant
figures if the value is above 10.

Report all pesticide/Aroclor results to two significant figures.  The
appropriate concentration units, ug/L or ug/kg, must be entered.

If the result is a value greater than or equal to the  quantitation
limit, report the value.

Under the column labeled "Q" for qualifier, flag each  result with  the
specific Data Reporting Qualifiers listed below.  The  Contractor is
encouraged to use additional flags or footnotes.  The  definition of
such flags must be explicit and must be included in  the SDG Narrative.

For reporting results to the USEPA,  the following contract specific
qualifiers are to be used.  The seven qualifiers defined below are not
subject to modification by the laboratory.  Up to five qualifiers may
be reported on Form I for each compound.
                             B-32                         OLM01.3  2/91

-------
The seven EPA-defined qualifiers to be used are as follows:

    U  -   Indicates compound was analyzed for but not detected.  The
         sample quantitation limit must be corrected for dilution and
         for percent moisture.  For example, 10 U for phenol  in water if
         the sample final volume is the protocol-specified final volume.
         If a 1 to 10 dilution of extract is necessary, the reported
         limit is 100 U.  For a soil sample, the value must also be
         adjusted for percent moisture.  For example, if the  sample had
         24% moisture and a 1 to 10 dilution factor, the sample
         quantitation limit for phenol (330 U) would be corrected to

             (330 U)  x df    where D = 100 - % moisture
                D                             100

         and df = dilution factor

         For example,  at 24% moisture, D = 100-24 =0.76
                                             100

             (330 U)  x 10 = 4300 U   rounded to the appropriate
                                      number of significant  figures
               .76

         For semivolatile soil samples, the extract must be concentrated
         to 0.5 mL, and the sensitivity of the analysis is not
         compromised by the cleanup procedures.  Similarly, pesticide
         samples subjected to GPC are concentrated to 5.0 mL.
        Therefore, the CRQL values in Exhibit C will apply to all
         samples,  regardless of cleanup.   However, if a sample extract
         cannot be concentrated to the protocol-specified volume (see
         Exhibit C),  this fact must be accounted for in reporting the
         sample quantitation limit.

   J  -   Indicates an estimated value.  This flag is used under the
        following circumstances: 1) when estimating a concentration for
         tentatively identified compounds where a 1:1 response is
        assumed,  2)  when the mass spectral and rentention time data
         indicate the  presence of a compound that meets the volatile and
        semivolatile  GC/MS identification criteria, and the  result is
        less than the CRQL but greater than zero, 3) when the retention
        time data indicate the presence of a compound that meets the
        pesticide/Aroclor identification criteria and the result is
        less than the CRQL but greater than zero.  Note:  the "J" code
        is not used and the compound is not reported as being
        identified for pesticide/Aroclor results less than the CRQL, if
        the technical judgement of the pesticide residue analysis
        expert determines that the peaks used for compound
        identification resulted from instrument noise or other
        interferences (column bleed,  solvent contamination,   etc.).   For
        example,  if the sample quantitation limit is 10 ug/L, but a
        concentration of 3 ug/L is calculated, report it as  3J.  The
        sample quantitation limit must be adjusted for dilution as
        discussed for the U flag.
                            B-33                         OLM01.8  8/91

-------
Indicates presumptive evidence of a compound.   This flag is
only used for tentatively identified compounds,  where the
identification is based on a mass spectral library search.   It
is applied to all TIC results.   For generic characterization of
a TIC, such as chlorinated hydrocarbon,  the N  code is not used.
                    B-33a                         OLM01.8  8/91

-------
 P -  This  flag  is used for a. pescicide/Aroclor target analyte when
     there  is greater than 25% difference for detected
     concentrations between the two GC columns (see Form X).  The
     lover  of the two values is reported on Form I and flagged with
     a "P".

 C -  This  flag  applies to pesticide results where the identification
     has been confirmed by GC/MS.  If GC/MS confirmation was
     attempted  but was unsuccessful, do not apply this flag, instead
    • use a  laboratory-defined flag, discussed below.

 B -  This  flag  is used when the analyte is found in the associated
     blank  as well as in the sample.  It indicates possible/probable
     blank  contamination and warns the data user to take appropriate
     action.  This flag must be used for a TIC as well as for a
     positively identified target compound.

 E -  This  flag  identifies compounds whose concentrations exceed the
     calibration range of the GC/MS instrument for that specific
     analysis.  If one or more compounds have a response greater
     than full  scale, except as noted in Exhibit D, the sample or
     extract must be diluted and re-analyzed according to the
     specifications in Exhibit D.  All such compounds with a
     response greater than full scale should have the concentration
     flagged with an "E" on the Form I for the original analysis.
     If the dilution of the extract causes any compounds identified
     in the first analysis to be below the calibration range in the
     second analysis, then the results of both analyses shall be
     reported on separate copies of Form I.  The Form I for the
     diluted sample shall have the "DL" suffix appended to the
     sample number.  NOTE:  For total xylenes, where three isomers
     are quantified as two peaks, the calibration range of each peak
     should be  considered separately, e.g., a diluted analysis is
     not required for total xylenes unless the concentration of the
     peak representing the single isomer exceeds 200 ug/L or the
     peak representing the two coeluting isomers on that GC column
     exceeds 400 ug/L.   Similarly, if the two 1,2-Dichloroethene
     isomers coelute, a diluted analysis is not required unless the
     concentration exceeds 400 ug/L.

D  -  This flag  identifies all compounds identified in an analysis ac
     a secondary dilution factor.  If a sample or extract is
     re-analyzed at a higher dilution factor, as in the "E" flag
     above, the "DL" suffix is appended to the sample number on the
     Form I for the diluted sample, and all concentration values
     reported on that Form I are flagged with the "D" flag.  This
     flag alerts data users that any discrepancies between the
     concentrations reported may be due to dilution of the sample or
     extract.

A  -  This flag  indicates that a TIC is a suspected
     aldol-condensation product.
                          B-34                         OLM01.2   1/91

-------
          X -  Other specific  flags  may be  required  to  properly  define  the
               results.   If used,  they must be  fully described,  and  such
               description attached  to the  Sample  Data  Summary Package  and  the
               SDG Narrative.   Begin by using "X".   If  more  than one flag is
               required,  use "Y"  and "Z"  as needed.  If more than five
               qualifiers are  required for  a sample  result,  use  the  "X" flag
               to combine several flags,  as needed.  For  instance, the  "X"
               flag might combine the "A",  "B", and  "D" flags for some
               samples.   The laboratory-defined flags are  limited to the
               letters "X",  "Y",  and "Z".

      The combination of flags "fill"  or "UB" is expressly prohibited.  Blank
      contaminants  are flagged "B"'only when they are detected  in the sample.

2.    Form  I VOA-TIC  and Form  I SV-TIC

      Fill  in all header information as above.

      Report Tentatively Identified  Compounds  (TICs) including  CAS  number,
      compound name,  retention time, and the estimated concentration
      (criteria  for reporting TICs are given in Exhibit D,  Section  IV).
      Retention  time  must be reported in minutes and decimal minutes, not
      seconds or minutes:seconds.

      If, in the opinion of the mass spectral  interpretation specialist,  no
      valid tentative identification can be made, the compound  shall be
      reported as unknown.

      Include a  Form  I VOA-TIC or SV-TIC for every volatile and semivolatile
      fraction of every  sample and method blank analyzed, even  if no TICs are
      found.  Total the  number of TICs found,   including aldol-condensation
      products (but see  below), and  enter  this number  in  the "Number TICs
      found." If none were found, enter "0" (zero).  Form I VOA-TIC or SV-TIC
      must  be provided for every analysis, including required dilutions and
      reanalyses, even if no TICs are found.

      If the name of  a compound exceeds the 28 spaces  in  the TIC column,
      truncate the  name  to 28 characters.  If  the compound  is an unknown,
      restrict description to no more than 28  characters  (i.e.,  unknown
      hydrocarbon,  etc.).

      Peaks that are  suspected as aldol-condensation reaction products (i.e.,
      4-methyl-4-hydroxy-2-pentanone and 4-methyl-3-pentene-2-one)  shall be
      summarized on this form, flagged "A", and included  in the total  "Number
      TICs  found,"  but not counted as part of  the 20 most intense non-target
      semivolatile  compounds to be searched.

C.    System Monitoring  Compound Recovery  (Form II VOA)

      For volatiles,  Form II is used to report the  recoveries of the system
      monitoring compounds added to  each volatile sample, blank, matrix
      spike, and matrix  spike  duplicate prior  to  analysis.   The system
      monitoring compounds, previously termed  volatile surrogates,  are used
      to monitor the  performance of  the purge  and trap-gas  chromatograph-mass
                                   B-35                         OLM01.3  2/91

-------
      spectrometer system as a whole.   Form II VOA is matrix-specific,  so
      that system monitoring compound recoveries for water samples are
      reported on a different version of Form II than the recoveries in soil
      samples.  Soil sample recoveries are further differentiated by
      concentration level.

      Complete the header information and enter EPA Sample Numbers as
      described in part A.  For volatile soil samples only, specify the
      "level" as "LOW" or "MED", as on Form I.  Do not mix low and medium
      level samples on one form.  Complete one for each level.  For each
      system monitoring compound, report the percent recovery to the nearest
      whole percentage point, and to the number of significant figures  given
      by the QC limits at the bottom of the form.

      Flag each system monitoring compound recovery outside the QC limits
      with an asterisk (*).  The asterisk must be placed in the last space in
      each appropriate column, under the "#" symbol.  In the far righthand
      column, total the number of system monitoring compound(s) recoveries
      outside the QC limits for each sample.  If no system monitoring
      compound(s) were outside the limits, enter "0".

      If the system monitoring compound(s) are diluted out in any analysis,
      enter the calculated recovery, or "0" (zero) if the system monitoring
      compound(s) is not detected, and flag the system monitoring compound(s)
      recoveries with a "D" in the column under the "#" symbol.  Do not
      include results flagged "D" in the total number of recoveries for each
      sample outside the QC limits.

      Number all pages as described in part A.

D.    Surrogate Recovery (Form II, SV and PEST)

      Form II is used to report the recoveries of the surrogate compounds
      added to each semivolatile and pesticide/Aroclor sample, blank, matrix
      spike,  and matrix spike duplicate.  For semivolatiles, Form II is
      matrix-specific as well as fraction-specific, so surrogate recoveries
      for semivolatile water samples are reported on a different version of
      Form II than semivolatile soil sample surrogate recoveries.

      Complete the header information and enter EPA Sample Numbers as
      described in part A.  For semivolatile soil samples only, specify the
      "level" as "LOW" or "MED", as on Form I.  Do not mix low and medium
      level samples on one form.  Complete one for each level.  For each
      surrogate, report the percent recovery to the nearest whole percentage
      point,  and to the number of significant figures given by the QC limits
      at the bottom of the form.

      Flag each surrogate recovery outside the QC limits with an asterisk
      (*).  The asterisk must be placed in the last space in each appropriate
      column, under the "#" symbol.  In the far righthand column, total the
      number of surrogate recoveries outside the QC limits for each sample.
      If no surrogates were outside the limits, enter "0".
                                   B-36                               OLM01.0

-------
      If the surrogates are diluted out in any analysis,  enter  the  calculated
      recovery, or "0" (zero) if the surrogate is  not detected,  and flag the
      surrogate recoveries with a "D" in the column under the  "#" symbol.   Do
      not include results flagged "D" in the total number of recoveries  for
      each sample outside the QC limits.

      The pesticide surrogate recoveries must be reported from both GC
      columns used for the analyses.  Therefore, identify each GC column in
      the header, entering the stationary phase under "GC Column"  (previously
      called GC Column ID), and the internal diameter (ID)  of the column in
      millimeters under "ID".  The assignment of columns  as "1"  and "2"  is
      left to the discretion of the laboratory if  the analyses  are  performed
      by simultaneous injection into*a GC containing two  columns.   If so
      analyzed, the assignment of "GC Column 1" and "GC Column 2" must be
      consistent across all the reporting forms.   If the  analysis is not
      performed by simultaneous injection, then the assignment of GC Column
      number should be based on the chronological  order of the  two  analyses.

      The pesticide surrogate recovery limits are  only advisory, but the
      Contractor must flag those recoveries outside the advisory QC limits  or
      diluted out, nonetheless.  The total number  outside the  QC limits
      includes all values, regardless of GC column.

      Number all pages as described in part A.

E.     Matrix Spike/Matrix Spike Duplicate Recovery (Form  III,  All  Fractions)

      This form is used to report the results of the analyses of a  matrix
      spike and matrix spike duplicate.  The form  is matrix-specific for
      volatiles and semivolatiles.

      Complete the header information as instructed in Part A,  including the
      EPA Sample Number for the matrix spike, without the suffixes  MS or MSD.

      For volatile and semivolatile soil samples,  specify "level"  as "LOW"  or
      "MED", as on Form I.  SDGs containing soil samples  at both levels
      require MS/MSD at each level, therefore, for soils, prepare  one form
      for each level.

      All water samples are "LOW".  Therefore, there is no MS/MSD  for "medium
      level waters", and none shall be reported.

      In the upper box in Form III, under "SPIKE ADDED",  enter the calculated
      concentration in ug/L or ug/Kg (according to the matrix) that results
      from adding each spiked compound to the aliquot chosen for the matrix
      spike (MS).  For instance, for base/neutral  compounds in medium level
      soils, if 50 ug of spike are added to 1 g of soil,  the resulting
      concentration is 50,000 ug/Kg.  Enter the "SAMPLE CONCENTRATION",   in
      similar units, of each spike compound detected in the original sample.
      If a spike compound was not detected during  the analysis of  the
      original sample, enter the sample result as   "0" (zero).  Under "MS
      CONCENTRATION", enter the actual concentration of each spike compound
      detected in the matrix spike aliquot.  Calculate the percent recovery
      of each spike compound in the matrix spike aliquot to the nearest  whole
      percent, according  to Exhibit D, and enter under "MS % REC".   Flag all
                                   B-37                         OLM01.3  2/91

-------
percent recoveries outside the QC limits with an asterisk (*).   The
asterisk must be placed in the last space of the percent recovery
column, under the "#" symbol.

For pesticide/Aroclor matrix spikes and matrix spike duplicates, the
concentration used for "MS CONCENTRATION" AND "MSD CONCENTRATION" must
be the concentration of the spiked analyte reported on Form I for those
analyses.  Of the two concentrations calculated for each
pesticide/Aroclor target compound, one on each GC column, the lower
concentration is reported on Form I, and both concentrations are
reported on Form X.  The lower concentration is also reported on Form
III and used in the calculation of spike recovery, even if that
concentration yields a recovery value that is outside the advisory QC
limits.

Complete the lower box on Form III in a similar fashion, using the
results of the analysis of the matrix spike duplicate (MSD) aliquot.
Calculate the relative percent difference (RPD) between the matrix
spike recovery and the matrix spike duplicate recovery, and enter this
value in the lower box under "% RPD".  Report the relative percent
difference to the nearest whole percent.  Compare the RPDs to the QC
limits given on the form, and flag each RPD outside the QC limits with
an asterisk (*) in the last space of the "% RPD" column, under the "#"
symbol.

Summarize the values outside the QC limits at the bottom of the page.
No further action is required by the laboratory.  Performance-based QC
limits will be generated and updated from recovery and RPD data.

Method Blank Summary  (Form IV, All Fractions)

This form summarizes the samples associated with each method blank
analysis.   A copy of the appropriate Form IV is required for each
blank.

Complete the header information on Form IV as described in Part A.  The
"EPA Sample No." entered in the box at the top of Form IV shall be the
same number entered on the Form I for the blank itself.

For volatile blanks, enter the "Instrument ID", "Date Analyzed" "Time
Analyzed", "GC Column", "ID", and "Heated Purge (Y/N)".  Volatile
samples analyzed by the same purging technique, i.e., ambient purge, or
heated purge,  may be reported together on the same Form IV, if the same
method blank applies to those samples.  Thus, water samples and medium
soil sample may be combined on a single form, if run with a single
blank.

For semivolatile blanks, enter the "Instrument ID", "Date Analyzed",
"Matrix" and "Level".  All water blanks are "LOW".  The "Time Analyzed"
shall be in military time.

For pesticide/Aroclor blanks, enter the method of extraction as "SEPF"
for separatory funnel, "SONG" for sonication, or "CONT" for continuous
liquid-liquid extraction.
                             B-38                        OLM01.1  12/90

-------
      For pesticide/Aroclor blanks, there is no differentiation between
      medium and low level soil samples, so no "Level" is entered on this
      form.

      For semivolatile and pesticide/Aroclor method blanks, enter the date of
      extraction of the blank.

      If the samples associated with pesticide/Aroclor blank are subjected to
      sulfur cleanup, then the blank must also be subjected to sulfur
      cleanup.  If sulfur cleanup is employed, enter "Y" in the "Sulfur
      Cleanup" field, else, enter "N".   If only some of the samples
      associated with the method blank are subjected to sulfur cleanup, a
      separate sulfur cleanup blank is required (see Exhibit D PEST).  If a
      separate sulfur cleanup blank is prepared,  complete one version of Form
      IV associating all the samples with the method blank, and a second
      version of Form IV listing only those samples associated with the
      separate sulfur cleanup blank.  Note:  Subjecting all samples
      associated with a method blank to sulfur cleanup avoids the need for
      two forms.

      Pesticide/Aroclor contaminants must meet the identification criteria in
      Exhibit D PEST, which requires analysis of the blank on two different
      GC Columns.  Therefore, enter the date, time and instrument ID of both
      analyses of the blank on the pesticide method blank summary.  The
      information on the two analyses is differentiated as Date Analyzed (1),
      Date Analyzed (2), etc.  If the analyses were run simultaneously, the
      order of reporting is not important, but must be consistent with the
      information reported on all other pesticide forms.  Otherwise, (1)
      shall be the first analysis, and (2) the second.  Identify the GC
      Column and internal diameter as described previously.

      Enter "Lab File ID" only if GC/MS confirmation was attempted.
      Otherwise, leave blank.

      For all three fractions, as appropriate, summarize the samples
      associated with a given method blank in the table below the header,
      entering EPA Sample Number and Lab Sample ID.  For volatiles, enter the
      Lab File ID and time of analysis of each sample.  For semivolatiles,
      enter Lab File ID and Date Analyzed.  For pesticides/Aroclors, enter
      the dates of both analyses as Date Analyzed (1) and Date Analyzed (2),
      as discussed above.

      Number all pages as described in part A.

G.    GC/MS Instrument Performance Check and Mass Calibration (Form V VOA and
      SV)

      This form is used to report the results of GC/MS instrument performance
      check (previously known as "Tuning") for volatiles and semivolatiles,
      and to summarize the date and time of analysis of samples, standards,
      blanks,  matrix spikes, and matrix spike duplicates associated with each
      analysis of the instrument performance check solution.
                                   B-39                               OLM01.0

-------
Complete  the header  information as in Instruction A.  Enter the "Lab
File  ID"  for the  injection containing the instrument performance check
solution  (BFB for volatiles, DFTPP for semivolatiles).  Enter the
"Instrument ID".  Enter the date and time of injection of the
instrument performance check solution.  Enter time as military time.
For volatiles, indicate the purging method by entering "Y" for heated
purge, and "N" for ambient temperature purge, as described previously.
Water samples and medium soil sample extracts may be reported on the
same  Form V if analyzed together, as a single calibration may be
applied to both sample types.

For each  ion listed  on the form, enter the percent relative abundance
in the righthand column.  Report relative abundances to the number of
significant figures  given for each ion in the ion abundance criteria
column.

Note  that for both BFB and DFTPP, one or more of the high mass ions may
exceed the abundance of the ion listed on the form as the nominal base
peak, m/z 95 for BFB, and m/z 198 for DFTPP.  Despite this possibility,
all ion abundances are to be normalized to the nominal base peaks
listed on Form V (see Exhibits D and E).

All relative abundances must be reported as a number.  If zero, enter
"0",  not  a dash or other non-numeric character.  Where parentheses
appear, compute the percentage of the ion abundance of the mass given
in the appropriate footnote, and enter that value in the parentheses.

In the lower half of the form, list all samples, standards, blanks,
matrix spikes, and matrix spike duplicates analyzed under that
instrument performance check in chronological order, by time of
analysis  (in military time).  Refer to Part A for specific instructions
for identifying standards and blanks.  Enter "EPA Sample No.", "Lab
Sample ID", "Lab File ID", "Date Analyzed", and "Time Analyzed" for all
standards, samples,  blanks, matrix spikes, and matrix spike duplicates.

The GC/MS instrument performance check solution must be analyzed again
twelve hours from the time of injection of the instrument performance
check solution (BFB or DFTPP) listed at the top of the form.  In order
to meet these requirements, a sample, standard, blank, matrix spike, or
matrix spike duplicate must be injected within twelve hours of the
injection of the instrument performance check solution.

Number all pages as described in Instruction A.
                             B-40                               OLM01.0

-------
 H.    GC/MS  Initial  Calibration Data  (Form VI VGA, SV-1, SV-2)

      After  a GC/MS  system has undergone an initial five-point  calibration
      at  the specific  concentration levels described in Exhibits D and E, and
      after  all  initial calibration criteria have been met, the laboratory
      must complete  and submit a Form VI for each volatile or semivolatile
      target compound  initial calibration performed which is relevant to the
      samples, blanks, matrix spikes, matrix spike duplicates in the SDG,
      regardless of  when  that calibration was performed.

      Complete all header information as in Part A.  Enter the "Case No." and
      "SDG No."  for  the current data  package, regardless of the original Case
      for which  the  initial calibration was performed.  Enter "Instrument ID"
      and the date(s)  of  the calibration.  If the calendar date changes
      during the calibration procedure, the inclusive dates should be given
      on Form VI.  Enter  the injection times of the first and last of the
      standards  analyzed  under "Calibration Times".

      For volatiles, enter Heated Purge, and GC Column, ID, as on Form V.

      Enter  the  "Lab File ID" for each of the five calibration standards
      injected.  Complete the response factor data for the five calibration
      points, and then calculate and  report the average relative response
      factor (RRF) for all target compounds.  For volatiles, report the
      response factors for the system monitoring compounds in the calibration
      standards.  For  semivolatiles,  report the response factors for all
      surrogate compounds in the calibration standards.  The laboratory must
      report the relative standard deviation (%RSD) for all compounds.

              %RSD - Standard Deviation  x  100
                           Mean
               Where,
                                               -  *)'
               Standard Deviation  -    	
                                              n-1
                                         n
1/2
               Where,

               X£  - each individual  value  used to  calculate  the mean

               x   - the  mean of n values

               n   - the  total number of values
   For Semivolatiles, eight compounds:  2,4-Dinitrophenol,
2,4,5-Trichlorophenol, 2-Nitroaniline, 3-Nitroaniline, 4-Nitroaniline,
4-Nitrophenol, 4,6-Dinitro-2-Methylphenol, and Pentachlorophenol will only
require a four-point initial calibration at 50, 80, 120, and 160 total
nanograms because detection at less than 50 nanograms per injection  is
difficult.  If a four-point calibration is performed  for these compounds,
leave RF20 blank.
                                   B-41                         OLM01.2  1/91

-------
 In order to be used for the analysis of samples or sample extracts, the
 volatile and semivolatile initial calibration must meet the acceptance
 criteria for relative response factors outlined in Exhibits D and E.
 The compounds for which criteria have been developed for minimum RRF
 and maximum %RSD are indicated on the form by an "*".   All other
 compounds must meet a minimum RRF of 0.010.

 GC/EC Initial Calibration Data (Form VI PEST-1, PEST-2, PEST-3, PEST-4)

 The initial calibration of pesticides and Aroclors involves the
 determination of retention times, retention time windows, and
 calibration factors.  For single component pesticide target compounds,
 these data are calculated from the analyses of the Individual Standard
 Mixtures A and B at three different concentration levels.  For the
 multicomponent target compounds, these data are calculated from a
 single point calibration.

 For the three analyses of Individual Standard Mixture A (low point, mid
 point, and high point), and the three analyses of Individual Standard
 Mixture B performed on each GC column during an initial calibration,
 complete one copy of Form VI for each GC column used.

 Complete the header information as above.  Enter the Instrument ID, GC
 Column, and ID as described previously.  Enter the dates of analysis of
 the first and last of the six standards on each form under "Date(s)
Analyzed".  Under "Level (x low)", enter the concentration of the low
point, mid point, and high point calibration standards as a multiplier
of the low point.  Therefore, for the low point, enter "1.0".  The
concentration of the mid point standard is specified in Exhibit D as
 four times the low point, therefore, enter "4.0" for "mid".  The high
point standard must be at least 16 times the low point, but may be
higher, if that value lies within the linear range of the instrument,
as specified in Exhibit D.  Therefore, enter the appropriate multiplier
 to the high point standard concentration to one decimal place.

 For each standard analyzed, enter the retention time of each applicable
analyte in minutes and decimal minutes, under the appropriate
 concentration level.  Calculate the mean retention time of each analyte
 from the three individual mixtures, and report it under "Mean RT".
 Calculate the retention time window for each analyte using the
 specifications in Exhibit D, and enter the lower limit of the window
under RT Window "From", and the upper limit of the window under "To".
The retention times of the surrogates are reported from the analyses of
 Individual Mixture A and the windows are only required to be calculated
for Individual Mixture A.

For the six analyses of the Individual Standard Mixtures, the
laboratory must also complete the calibration factor data on Form VI
PEST-2.  In a similar fashion as for the retention time data on Form VI
PEST-1, prepare one form for each instrument and GC column used.  Enter
the concentration level of the standards in the same fashion as on  Form
VI PEST-1.
                             B-42                         OLM01.2  1/91

-------
 Enter  the  calibration  factor  for each compound in each of the
 standards,  and calculate a mean calibration and a percent relative
 standard deviation  (%RSD), and enter on the form.  As with surrogate
 retention  times,  the calibration factors are only required from
 Individual Mixture  A analyses.

 In order to be used for sample analyses, the- %RSD of the calibration
 factors 'for each  single component target compound must be less than or
 equal  to 20.0 percent  except  as noted in the following.  The %RSD of
 the calibration factors for the two surrogates must be less than or
 equal  to 30.0 percent.  Up to two single component target compounds
 (but not surrogates) may exceed the 20.0 percent limit for %RSD, but
 these  compounds must have a %RSD of less than of equal to 30.0 percent.
 These  criteria apply to both  GC columns.

 For the multicomponent target compounds, the retention times, retention
 time windows, and calibration factor must be reported in a similar
 fashion for each  single point calibration standard.  For each
 multicomponent compound, the  laboratory must select at least three
 peaks  from each analyte, according to the specifications in Exhibit D.
 The retention and calibration factor data apply to each peak.  Complete
 one version of Form VI PEST-3 for each GC column, for each initial
 calibration that  applies to samples in the data package.

 Form VI is  used also to report the results of analysis of the
 Resolution Check  Solution that must begin each pesticide/Aroclor
 initial calibration sequence.  The purpose of the Resolution Check
 Solution is to demonstrate for each initial calibration that the GC
 columns employed  are capable of satisfactorily resolving the most
 difficult  of the  target analytes.  One copy of Form VI PEST-4-is
 completed  that covers  both GC columns.

 Complete the header information as described in Instruction A.  Using
 the  same assignment of first and second GC columns made for Form IV,
 enter  the GC Column, ID, Instrument ID, and Date and Time Analyzed.
 Enter  the  "EPA Sample No." for the Resolution Check Standard.  If
 simultaneous injections on a single GC are used, the EPA Sample No. may
be  the same  for both Resolution Check Standards.  If simultaneous
 injections were not used, use different suffixes to identify the
 standards.

 In  the boxes on the form, list each analyte, in retention time order.
 including both surrogate compounds.  Thus, the order of analytes in the
 two boxes on a copy of this form will be different, due to the
dissimilarity of  the stationary phases of the two GC columns used.
Enter  the name of each target analyte in the Resolution Check Mixture
as  it  appears on  Form  I PEST.  Spell out the names of the surrogates as
they appear  on Form VII PEST-2.

Enter  the retention time of each analyte from the analysis under "RT".
Calculate the resolution between each pair of analytes according to the
definition  in Exhibit D PEST.  The resolution is calculated as
percentage of the height of the smaller of each pair of adjacent peaks.
Enter  the resolution between  the first and second peaks on the line for
                             B-43                         OLM01.2  1/91

-------
      the first analyte listed in the box.  Enter the resolution between the
      second and third peaks on the line for the second analyte,  and so on,
      until the resolutions of all possible pairs of adjacent analytes have
      been entered.  NOTE:  Only eight of the nine resolution fields will be
      filled.   In order for these GC columns to be used for pesticide/Aroclor
      analyses, the resolution of all pairs of peaks listed on this form must
      be greater than or equal to 60.0%.

J.    GC/MS Continuing Calibration Data (Form VII VOA, SV-1, SV-2)

      For volatiles and semivolatiles, the Continuing Calibration Check form
      is used to report the calibration of the GC/MS system by the analysis
      of specific calibration standards.  A Continuing Calibration Check form
      is required for each twelve (12) hour time period for both volatile and
      semivolatile target compound analyses.

      The Contract laboratory must analyze calibration standards and meet all
      criteria outlined in Exhibits D and E for the minimum RRF and maximum
      percent difference between initial and continuing calibrations.

      Complete all header information as in Instruction A.   Enter instrument
      ID, date and time of continuing calibration, the Lab  File ID of the
      continuing calibration standard, and dates and times  of initial
      calibration (give inclusive dates if initial calibration is performed
      over more than one date).   For volatiles, enter purge method and column
      as on Forms IV,  V and VI.   Using the appropriate Initial Calibration
      (volatile or semivolatile)  fill in the average relative response factor
      (RRF) for each target compound,  each system monitoring compound for
      volatiles,  and each surrogate for semivolatiles.  Report the relative
      response factor (RRF50)  from the continuing calibration standard
      analysis.   Calculate the Percent Difference (%D) for  all compounds.
                                -  RRFC
         % Difference  - - x 100
      Where
                 Average relative response factor from initial calibration.

         RRFC -  Relative response factor from continuing calibration
                   standard.

      All  semivolatile  standards  are  analyzed at  50  total ng.
                                   B-44                        OLM01.2   1/91

-------
K.    GC/EC Calibration Verfication Summary (Form VII PEST)

      The Calibration Verification Summary Form VII is used  to report the
      results of the Performance Evaluation Mixtures (PEM),  instrument
      blanks, and Individual Standard Mixtures A and B analyzed at the
      beginning and end of a twelve hour sequence.  The laboratory must
      submit this form for each twelve hour sequence analyzed.

      Complete the header information on each Form VII required according to
      the instructions in part A.

      Enter the initial calibration date(s) analyzed.  Give  inclusive dates
      if initial calibration is performed over more than one date.

      On Form VII PEST-1, enter the EPA Sample No., Lab Sample ID,-Date
      Analyzed, and Time Analyzed for the instrument blank that preceded the
      twelve hour sequence (PIBLK).   For the PEM that initiated or terminated
      the twelve hour sequence (PEM), enter the EPA Sample No., Lab Sample
      ID, Date Analyzed, and Time Analyzed.

      When reporting data for the PEM at the beginning of the initial
      calibration sequence,  leave blank the "EPA Sample No.", "Lab Sample
      ID", and "Date" and "Time Analyzed" fields for the instrument blank
      (PIBLK), as no instrument blank is analyzed before this PEM.  When
      reporting all other PEM analyses, the instrument blank fields must be
      completed.

      In the table,  report the retention time for each analyte in the PEM as
      well as the retention time windows.  For each analyte  in the PEM, enter
      the amount of the analyte calculated to be in the PEM, in nanograms to
      three decimal places,  under "CALC AMOUNT".  Enter the  nominal amount of
      each analyte in the PEM under "NOM AMOUNT".  Calculate the relative
      percent difference between the calculated amount and nominal amount for
      each analyte according to Exhibit D.  Report the values under "RPD".

      Calculate the percent breakdown for endrin and 4,4'-DDT, and the
      combined percent breakdown in the PEM according to Exhibit D.  Enter
      the values for the breakdown of endrin and 4,4'-DDT in their respective
      fields immediately under the table.

      Form VII PEST-2 is used to report the results of the analyses of the
      instrument blank and the mid point concentrations of Individual
      Standard Mixtures A and B that, along with the PEM, bracket each 12-
      hour period of sample analyses.  One copy of Form VII  PEST-2 must be
      completed each time the Individual Standard Mixtures are analyzed, for
      each GC column used.  The form is completed in a fashion similar to
      Form VII, entering the EPA Sample No., Lab Sample ID,  Date Analyzed,
      and Time Analyzed for the instrument blank immediately preceding the
      Individual Standard Mixtures A and B, and for the standards themselves.
      The upper table on the form contains the retention time and amount data
      for Individual Standard Mixture A compounds.  The lower table contains
      the data for Mixture B.  Enter the data in these tables in a fashion
      similar to that for the PEM.  Complete copies of Form VII PEST-1 and
      PEST-2 for each standard reported on Form VIII PEST.
                                   B-45                         OLM01.2  1/91

-------
L.    Internal Standard Area and RT Summary (Form VIII  VOA and SV)

      This form is used to summarize the  peak areas  and retention times  of
      the internal standards added to all volatile and  semivolatile samples,
      blanks,  matrix spikes, and matrix spike duplicates.   The data are  used
      to determine when changes in internal standard responses will adversely
      affect quantification of target compounds.   -This  form must be completed
      each time a continuing calibration  is performed,  or  when samples are
      analyzed under the same GC/MS instrument performance check as an
      initial calibration.

      Complete the header information as  in Instruction A.   Enter the Lab
      File ID of the continuing calibration standard, as well  as the date and
      time of analysis of the continuing  calibration standard.   If samples
      are analyzed immediately following  an initial  calibration, before
      another instrument performance check and a  continuing calibration,  Form
      VIII shall be completed on the basis of the internal standard areas of
      the 50 ug/L initial calibration standard for volatiles,  and the  50 ng
      initial calibration standard for semivolatiles.   Use the date and  time
      of analysis of this standard,  and its Lab File ID and areas in place of
      those of a continuing calibration standard.

      For volatiles, enter purge method and column,  as  on  Forms IV, V, VI,
      and VII.

      From the results of the analysis of the continuing calibration
      standard,  enter the area measured for each  internal  standard and  its
      retention time (in decimal minutes) under the  appropriate column  in the
      row labeled "12 HOUR STD".  For each internal  standard,  calculate  the
      upper limit of the area as the area of the  particular standard plus
      100% of its area (i.e., two times the area  in  the 12 HOUR STD box), and
      the lower limit of the area as the  area of  the internal  standard minus
      50% of its area (i.e., one half the area in the 12 HOUR  STD box).
      Report these values in the boxes labeled "UPPER LIMIT" and "LOWER
      LIMIT" respectively.  Calculate the upper limit of the retention  time
      as the retention of the internal standard plus 0.50  minutes (30
      seconds),  and the lower limit of the retention time  as the retention
      time in the standard minus 0.50 minutes (30 seconds).

      For each sample,  blank, matrix spike, and matrix  spike duplicate
      analyzed under a given continuing calibration, enter the EPA Sample
      Number and the area measured for each internal standard  and its
      retention time.   If the internal standard area is outside the upper or
      lower limits calculated above, flag that area  with an asterisk (*).
      The asterisk must be placed in the  far right hand space  of the box for
      each internal standard area, directly under the "#"  symbol.  Similarly,
      flag the retention time of any internal standard  that is outside  the
      limits with an asterisk.

      Number all pages as described in Instruction A.
                                   B-46                               OLM01.0

-------
M.    Pesticide Analytical Sequence (Form VIII Pest)

      This form is used to report the analytical sequence for pesticide
      analysis.  At least one Form VIII PEST is required for each GC column
      used for pesticide/Aroclor analyses.

      The laboratory shall complete all the header information as in Part A.
      Enter dates of analyses for the initial calibration, GC column, ID, and
      Instrument ID, as on Forms IV, VI, and VII.

      At the top of the table, report the mean retention time for
      tetrachloro-m-xylene and decachlorobiphenyl calculated from the initial
      calibration sequence under "TCX" and "DCB", respectively.  For every
      analysis associated with a particular analytical sequence starting with
      the initial calibration, enter the EPA Sample Number, Lab Sample ID,
      Date Analyzed, and Time Analyzed.  Each sample analyzed as part of the
      sequence must be reported on Form VIII PEST even if it is not
      associated with the SDG.  The laboratory may use the EPA Sample No. of
      "ZZZZZ" to distinguish all samples that are not part of the SDG being
      reported.  Report the retention time of the surrogates for each
      analysis under "TCX RT" and "DCB RT".  All sample analyses must be
      bracketed by acceptable analyses of instrument blanks, a PEM, and
      Individual Standard Mixtures A and B.  Given the fact that the initial
      calibration may remain valid for some time (see Exhibit D), it is not
      necessary to report the data from 12-hour periods when no samples in an
      SDG were run.  The laboratory must deliver the Form VIII for the
      initial calibration sequence, and Forms that include the PEMs and
      Individual Standard Mixtures that bracket any and all samples in the
      SDG.  While the data for time periods between the initial calibration
      and samples in the SDG is not a routine deliverable, it must be made
      available on request during on-site evaluations, etc.  Here again, non-
      EPA samples may be indicated with "ZZZZZ".

      Flag all those values which do not meet the contract requirements by
      entering an asterisk (*) in the last column, under the "*".  If the
      retention time cannot be calculated due to interfering peaks, leave the
      RT column blank for that surrogate, enter an asterisk in the last
      column, and document the problem in the SDG Narrative.

      If more than a single copy of Form VIII PEST is required, enter the
      same header information on all subsequent pages for that GC Column and
      Instrument, and number each page as described in Part A.

      Form VIII PEST is required for each for each GC system and for each GC
      column used to analyze target pesticides/Aroclors.

N.    Pesticide Cleanup Summary (Form IX PEST-1, PEST-2)

      This form summarizes the results of the checks performed for both
      cleanup procedures employed during the preparation of pesticide
      extracts for analysis.  Form IX PEST-1 is used to report the results of
                                   B-47                               OLM01.0

-------
Che check of the Florisil cartridges used to process all sample
extracts, and to associate the lot of cartridges with particular sample
results.  In this fashion, problems with a lot of cartridges may be
tracked across many samples.

Complete the header information on each Form IX required, according to
the instructions in Part A.

Enter the "Case No." and "SDG No." for the current data pa.ckage,
regardless of the original Case for which the cartridge check was
performed.  Enter the "Florisil Cartridge Lot Number".  Enter under the
"Date Analyzed" the date the Florisil cartridge check solution was
analyzed.

Enter "GC Column" and "ID" for the GC column used to determine the
recovery of the analytes in the Florisil cartridge check solution,
under "GC Column (1)".  If more than one GC column is used, enter the
information for a second column under "GC Column (2)", etc., as
discussed previously.

In the upper table, enter the amount of spike added and spike recovered
in nanograms for each analyte.

Calculate to the nearest whole percent, and enter the percent recovery
in the "% REC" field.  Flag each spike recovery outside the QC limits
with an asterisk (*).  The asterisk must be placed in the last space in
the "% REC" column, under the "#" symbol.

In the lower table, enter the "EPA Sample No.", the "Lab Sample ID",
and "Date Analyzed" for each sample and blank that was cleaned up using
this lot of Florisil cartridges.

Number the Form IX pages as described in Part A.

Form IX PEST-2 summarizes the results of the calibration of the Gel
Permeation Chromatography device (GPC) that must be used to process all
soil sample extracts for pesticide/Aroclor analyses.

Complete all header information as in Part A.  Enter an identifier for
the GPC Column, and the date of calibration in the appropriate fields.
Enter the two "GC Column" and "ID" fields, as discussed above.

For each of the pesticide matrix spike compounds listed in the box in
the upper portion of the form, enter the amount of the spike added to
the GPC column in ng, and the amount recovered, also in ng.  Calculate
the percent recovery of each analyte, and enter these values on the
form,  to the nearest percent.   Compare the recoveries to the QC limits
shown on the form,  and flag all those values outside the limits with an
asterisk (*) in the column under the "#" symbol.
                             B-48                         OLM01.2  1/91

-------
      For each sample in the data package that was subjected to GPC under
      this calibration, enter the EPA Sample No.,  Lab Sample ID,  and the date
      of both analyses in the lower portion of the form.

      If more than one copy of Form IX PEST-2 is required,  number all pages
      as described in Instruction A.

0.    Pesticide/Aroclor Identification (Form X PEST-1,  PEST-2)

      This form summarizes the quantitations of all target
      pesticides/Aroclors detected in a given sample.  It reports the
      retention times of the compound on both columns on which  it was
      analyzed,  as well as the retention time windows of the standard for
      that compound on both of these columns.  In addition,  it  is used to
      report the concentration determined from each GC column,  and the
      percent difference between the two quantitative results.   Separate
      copies of Form X are used for single component analytes and
      multicomponent analytes.

      Copies of Form X are required for each sample, blank,  matrix spike, and
      matrix spike duplicate in which target pesticides or Aroclors are
      detected.   If none are detected in a given sample,  no copy of Form X is
      required for that sample.

      Complete the header information as in Instruction A.   Enter the GC
      Column,  and ID for each of the two columns,  one as GC Column (1), the
      other as (2),  as described previously.  Enter the Instrument ID
      associated with each GC column directly below.

      For each single component pesticide detected, enter the name of the
      compound under "ANALYTE" as it appears on Form I.  Enter  the retention
      times on each column of the compounds detected in the sample next to
      the appropriate column designation (1 or 2).  Enter the retention time
      windows  on each column from the initial calibration standard.  These
      data must correspond with those on Form VI,  and are entered in a
      similar  manner.  The lower value is entered under the "FROM" column,
      the upper value under the "TO" column.

      Enter the concentration calculated from each GC column under the column
      labeled  "CONCENTRATION".  The units are the  same as those used .on Form
      I,  ug/L  for water samples,  and ug/Kg for soil samples. However, do not
      enter any units on Form X.   Calculate the percent difference between
      the concentrations entered on this form, using the equation below, and
      report it to a tenth of a percent under "%D".

                              %D - ConcH -  ConcL
                                        '	  x 100
                                   B-49                        OLM01.2   1/91

-------
Where,

Concn - The higher of the two concentraiton for the target compound in
question

Conci. - The lower of the two concentrations for the target compound in
question

Note that using this equation will result in percent difference values
that are always positive.  The value will also be greater than a value
calculated using the higher concentration in the denominator, however,
given the likelihood of a positive interference raising the
concentration determined on one GC column, this is a conservative
approach to comparing the two concentrations.

The lower of the two concentrations is reported on Form I for each
pesticide compound.  The lower concentration is used because, if
present, co-eluting interferences are likely to increase the calculated
concentration of any target compound.  If the percent difference
between the calculated concentrations is greater than 25.0 percent,
flag the concentration on Form I, as described previously.  This will
alert the data user to the potential problems in quantitating this
analyte.

If more pesticide compounds are identified in an individual sample than
can be reported on one copy of Form X, then complete as many additional
copies of Form X as necessary, duplicating all header information, and
numbering the pages as described in Instruction A.

Multicomponent analytes detected in samples are reported on a separate
version of Form X.  Complete the header information and Instrument and
GC Column fields as described above.  For multicomponent analytes, it
is necessary to report the retention time and concentration of each
peak chosen for quantitation in the target analyte, in fashion similar
to that for single component pesticides.  The concentrations of all
peaks quantitated (three are required, up to five may be used) are
averaged to determine the mean concentration.   Report the lower of the
two mean concentrations on Form I.   Flag this value as described
previously, if the mean concentrations from the two GC columns differ
by more than 25.0 percent.

If more multicomponent compounds are identified in an individual sample
than can be reported on one copy of Form X, then complete as many
additional copies of Form X as necessary, duplicating all header
information, and numbering the pages as described in Instruction A.
                             B-50                         OLM01.2  1/91

-------
                   THE FOLLOWING ARE DOCUMENT  CONTROL  FORMS
                      (To be  submitted as  hardcopy  only)

P.     Sample Log-In Sheet (Form DC-1)

      This form is used to document the receipt and inspection of sample
      containers and samples.  One original of Form DC-1  is  required for each
      sample shipping container.  If the samples in a  single sample shipping
      container must be assigned to more than  one Sample  Delivery Group, the
      original Form DC-1 shall be placed with  the deliverables for the Sample
      Delivery Group of the lowest Arabic number and a copy of Form DC-1
      must be placed with the deliverables for the  other  Sample Delivery
      Group(s).  The copies should be identified as "copy(ies)," and the
      location of the original should be noted on the  copies.

      Sign and date the airbill (if present).   Examine the shipping container
      and record the presence/absence of custody seals and their condition
      (i.e., intact, broken)  in item 1 on Form DC-1.   Record the custody seal
      numbers in item 2.

      Open the container, remove the enclosed  sample documentation, and
      record the presence/absence of chain-of-custody  record(s), SMO forms
      (i.e., Traffic Reports, Packing Lists),  and airbills or airbill
      stickers in items 3-5 on Form DC-1.   Specify  if  there  is an airbill
      present or an airbill sticker in item 5  on Form  DC-1.   Record the
      airbill or sticker number in item 6.

      Remove the samples from the shipping container(s),  examine the samples
      and the sample tags (if present),  and record  the condition of the
      sample bottles (i.e., intact, broken, leaking) and  presence of absence
      of sample tags in items 7 and 8 on Form  DC-1.

      Review the sample shipping documents and complete the  header
      information described in Instruction A.   Compare the information
      recorded on all the documents and samples and circle the appropriate
      answer in item 9 on Form DC-1.

      If there are no problems observed during receipt, sign and date
      (include time) Form DC-1, the chain-of-custody record, and Traffic
      Report, and write the sample numbers on  Form  DC-1.   Record the
      appropriate sample tags and assigned laboratory  numbers if applicable.
      The log-in date should be recorded at the top of Form DC-1 and the date
      and time of cooler receipt at the laboratory  should be recorded in
      items 10 and 11.   Record the fraction designation (if appropriate) and
      the specific area designation (e.g., refrigerator number) in the Sample
      Transfer block located in the bottom left corner of Form DC-1.  Sign
      and date the Sample Transfer block.   Cross out unused columns and
      spaces.
                                   B-51                               OLM01.0

-------
      If there are problems observed during receipt or an answer marked with
      an asterisk (i.e., "absent*") was circled, contact SMO and document the
      contact as well as resolution of the problem on a CLP Communication
      Log.  Following resolution, sign and date the forms as specified in the
      preceding paragraph and note, where appropriate, the resolution of the
      problem.

Q.    Document Inventory Sheet (Form DC-2)

      This form is used to record the inventory of the SDG File Purge
      documents and count of documents in the original Sample Data Package
      which is sent to the Region.

      Organize all EPA-CSF documents as described in Exhibit B, Section II
      and Section III.  Assemble the documents in the order specified on Form
      DC-2 and Section II and III, and stamp each page with a consecutive
      number.  (Do not number the DC-2 form).  Inventory the CSF by reviewing
      the document numbers and recording page numbers ranges in the columns
      provided in the Form DC-2.  If there are no documents for a specific
      document type, enter an "NA" in the empty space.

      Certain laboratory specific documents related to the CSF may not fit
      into a clearly defined category.  The laboratory should review DC-2 to
      determine if it is most appropriate to place them under No. 7, 8, 9, or
      10.  Category 10 should be used if there is no appropriate previous
      category.  These types of documents should be described or listed in
      the blanks under each appropriate category.
                                   B-52                               OLM01.0

-------
     SECTION IV






DATA REPORTING FORMS
     B'53                                OLM01.0

-------
                             1A
              VOLATILE ORGANICS ANALYSIS DATA SHEET
                                         EPA SAMPLE NO.
Lab Name:

Lab Code:
Case No.:
Contract:

 SAS No.:
SDG No.:
Matrix: (soil/water)	

Sample wt/vol:      	(g/mL)

Level:   (low/med)  	

% Moisture: not dec.	

GC Column:  	 ID: 	

Soil Extract Volume: 	(uL)
                          Lab Sample ID:

                          Lab File ID:

                          Date Received:

                          Date Analyzed:
           (mm)
         Dilution Factor:
       CAS NO.
   COMPOUND
         Soil Aliquot Volume:

   CONCENTRATION UNITS:
   (ug/L or ug/Kg)	
                                                    .(uL)
       74-87-3	Chloromethane	
       74-83-9	Bromomethane	
       75-01-4	Vinyl Chloride	
       75-00-3	Chloroe thane	
       75-09-2	Methylene Chloride	
       67-64-1	Acetone	
       75-15-0	Carbon Disulfide	
       75-35-4	1,1-Dichloroethene	
       75-34-3	1,1-Dichloroethane	
       540-59-0	1,2-Dichloroethene (total)
       67-66-3	Chloroform	[
       107-06-2	1,2-Dichloroethane	
       78-93-3	2-Butanone    	
       71-55-6	1,1,1-Trichloroethane	
       56-23-5	Carbon Tetrachloride	
       75-27-4	Bromodichlorome thane	
       78-87-5	1, 2-Dichloropropane	
       10061-01-5	cis-1,3-Dichloropropene	
       79-01-6	Trichloroethene	
       124-48-1	Dibromochlorome thane	
       79-00-5	1,1, 2-Trichloroethane	
       71-43-2	Benzene	
       10061-02-6	trans-1, 3-Dichloroproper.e_
       75-25-2	Bromoforni	
       108-10-1	4-Methyl-2-Per.tanone	
       591-73-6	2-Hexanone	
       127-18-4	Tetrachloroethene	
       79-34-5	1, l,2,2-Tetrachlorocthane_
       108-88-3	Toluene	
       108-90-7	Chlorobcnzeno	
       100-41-4	Ethylbenzenc	~	
       100-42-5	Styrene	'_	'
       1330-20-7	Xyler.e  (total)	  	
                                FORM I VGA
                                                  3/90

-------
                             IB
          SEMIVOLATILE ORGANICS ANALYSIS DATA SHEET
                                         EPA SAMPLE NO.
Lab Name:

Lab Code:
Case No.:
Contract:

 SAS No.:
SDG No.:
Matrix:  (soil/water)	

Sample wt/vol:      	(g/mL)	

Level:    (low/med)  	

% Moisture:  	  decanted: (Y/N)	

Concentrated Extract Volume:  	(uL)

Injection Volume:   	(uL)

GPC Cleanup:   (Y/N)	       pH:	
                          Lab Sample ID:

                          Lab File ID:

                          Date Received:

                          Date Extracted:

                          Date Analyzed:
                          Dilution Factor:
       CAS NO.
   COMPOUND
   CONCENTRATION UNITS:
   (ug/L or ug/Kg)	
            Q
       108-95-2	Phenol	
       111-44-4	bis (2-Chloroethyl) ether	
       95-57-8	2-Chlorophenol	
       541-73-1	1, 3-Dichlorobenzene	
       106-46-7	1, 4-Dichlorobenzene	
       95-50-1	1, 2-Dichlorobenzene	
       95-48-7	2-Methylphenol	
       108-60-1	2, 2 ' -oxybis (1-Chloropropane)
       106-44-5	4-Methylphenol	
       621-64-7	N-Nitroso-di-n-propylamine	
       67-72-1	Hexachloroethane	
       98-95-3	:	Nitrobenzene	
       78-59-1	Isophorone	
       88-75-5	2-Nitrophenol	
       105-67-9	2, 4-Dimethylphenol	
       111-91-1-	bis (2-Chloroethoxy) methane	
       120-83-2	2,4-Dichlorophenol	
       120-82-1	1,2,4-Trichlorobenzene	
       91-20-3	Naphthalene	
       106-47-8	4-Chloroaniline	
       8.7-68-3	Hexachlorobutadiene	
       59-50-7	4-Chloro-3-methyl phenol	
       91-57-6	2-Methylnaphthalene	
       77-47-4	Hexachlorocyclopentadiene	
       88-06-2	2,4 , 6-Trichlorophenol	
       95-95-4	2 , 4 , 5-Trichlorophenol	
       91-58-7	2-Chloronaphthalene	
       88-74-4	2-Nitroaniline	
       131-11-3	Dimethylphthalate	
       208-96-8	Acenaphthylene	
       606-20-2	2, 6-Dinitrotoluene	
       99-09-2	3-Nitroaniline	
       83-32-9	Acenaphthene	
                                FORM I SV-1
                                                  3/90

-------
                             1C
          SEMIVOLATILE ORGANICS ANALYSIS DATA SHEET
                                                             EPA SAMPLE NO.
 •b Name:

Lab Code:
Case No.:
                                     Contract:

                                      SAS No. :
                                    SDG No.:
                          .(g/iuL).
Matrix: (soil/water)	

Sample wt/vol:      	

Level:   (low/med)  	

% Moisture:  	  decanted: (Y/N)

Concentrated Extract Volume: 	

Injection Volume:   	(uL)

GPC Cleanup:   (Y/N)	       pH:
                                    (uL)
                          Lab Sample ID:

                          Lab File ID:

                          Date Received:

                          Date Extracted:

                          Date Analyzed:
                                              Dilution Factor:
       CAS NO.
                       COMPOUND
                    CONCENTRATION UNITS:
                    (ug/L or ug/Kg)	
       51-28-5	2 , 4-Dinitrophenol   	
       100-02-7	4-Nitrophenol	
       132-64-9	Dibenzofuran	
       121-14-2	2, 4-Dinitrotoluene	
       84-66-2	Diethylphthalate	
       7005-72-3	4-Chlorophenyl-phenylether_
       86-73-7	Fluorene	
       100-01-6	4-Nitroaniline	
       534-52-1	4 , 6-Dinitro-2-methylphenol_
       86-30-6	N-Nitrosodiphenylamine  (1)_
       101-55-3—	4-Bromophenyl-phenylether_2

       87-86-5	Pentachlorophenol
       85-01-8	Phenanthrene	
       120-12-7	Anthracene	
       86-74-8	Carbazole	
       84-74-2	Di-n-butylphthalate	
       206-44-0	Fluoranthene	
       129-00-0	Pyrene	
       85-68-7	Butylbenzylphthalate	
       91-94-1	3 , 3 ' -Dichlorobenzidine	
       56-55-3	Benzo (a) anthracene	
       218-01-9	Chrysene	
       117-81-7	bis(2-Ethylhexyl)phthalate_
       117-84-0	Di-n-octylphthalate	
       205-99-2	Benzo(b) fluoranthene	
       207-08-9	Benzo(k) f luoranthene	
       50-32-8	Benzo (a) pyrene	
       193-39-5	Indeno (1,2 , 3-cd) pyrene	
       53-70-3	Dibenz(a,h) anthracene	
       191-24-2	Benzo (g,h,i) pei
    (1) - Cannot be  separated  from  Diphenylamine

                                FORM I SV-2
                                                                      3/90

-------
                              ID
              PESTICIDE  ORGANICS ANALYSIS DATA SHEET
                                         EPA SAMPLE NO.
Lab Name:

Lab Code:
Case No.:
Contract:

 SAS No.:
Matrix:  (soil/water)	

Sample wt/vol:      	(g/mL)	

% Moisture:  	  decanted:  (Y/N)
Extraction:  (SepF/Cont/Sonc)

Concentrated Extract Volume: 	

Injection Volume:   	(uL)

GPC Cleanup:   (Y/N)	       pH:
                (uL)
SDG No.:
         Lab Sample ID:

         Lab File ID:

         Date Received:

         Date Extracted:

         Date Analyzed:
                          Dilution Factor:
                          Sulfur Cleanup:  (Y/N)
       CAS NO.
   COMPOUND
   CONCENTRATION UNITS:
   (ug/L or ug/Kg)	
       319-84-6	alpha-BHC	
       319-85-7	beta-BHC	
       319-86-8	delta-BHC	
       58-89-9	gamma-BHC (Lindane)
       76-44-8	•	Heptachlor	\
       309-00^-2	Aldrin	
       1024-57-3	Heptachlor epoxide_
       959-98-8	-Endosulfan I	
       60-57-1	Dieldrin	
       72-55-9	4 , 4 • -DDE	
       72-20-8	Endrin	
       33213-65-9	Endosulfan II	
       72-54-8	4 , 4 ' -ODD	
       1031-07-8	Endosulfan sulfate_
       50-29-3	4 , 4 • -DDT	
       72-43-5	Methoxychlor	
       53494-70-5	Endrin ketone	
       7421-36-3	Endrin aldehyde	
       5103-71-9	alpha-Chlordane	
       5103-74-2-	gamma-Chlordane	
       8001-35-2	Toxaphene	
       12674-11-2	Aroclor-1016	
       11104-28-2	Aroclor-1221	
       11141-16-5	Aroclor-1232	
       53469-21-9	Aroclor-1242	
       12672-29-6	Aroclor-1248	
       11097-69-1	Aroclor-1254	
       11096-82-5	Aroclor-1260
                                FORM I  PEST
                                                  3/90

-------
Lab Name:

Lab Code:
                              IE
              VOLATILE ORGANICS ANALYSIS DATA SHEET
                TENTATIVELY  IDENTIFIED COMPOUNDS
      Contract:

       SAS No.:
                              EPA SAMPLE NO.
Case No.:
SDG No.:
Matrix:  (soil/water)	

Sample wt/vol:      	(g/mL)

Level:    (low/med)  	

% Moisture: not dec.	

GC Column: 	 ID: 	

Soil Extract Volume: 	(uL)
               Lab Sample ID:

               Lab File ID:

               Date Received:

               Date Analyzed:
(mm)
                          Dilution Factor:
 Number TICs found:
               Soil  Aliquot Volume:

         CONCENTRATION UNITS:
         (ug/L or  ug/Kg)	
                                                    (uL)
CAS NUMBER
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.
20.
21.
22.
23.
24.
25.
26.
27.
28.
29.
30.

COMPOUND NAME































RT































EST . CONC .































Q































                                FORM I VOA-TIC
                                       3/90

-------
Lab Name:

Lab Code:
                              IF
          SEMIVOLATILE  ORGANICS  ANALYSIS  DATA  SHEET
                TENTATIVELY  IDENTIFIED  COMPOUNDS
Contract:

 SAS No.':
                        EPA SAMPLE NO.
Case No.:
                                    SDG No.:
Matrix:  (soil/water)	

Sample wt/vol:      	(g/mL)	

Level:    (low/med)  	

% Moisture:  	  decanted:  (Y/N)	

Concentrated Extract Volume: 	(uL)

Injection Volume:   	(uL)

GPC Cleanup:    (Y/N)	       pH:	
         Lab Sample ID:

         Lab File ID:

         Date Received:

         Date Extracted:

         Date Analyzed:
         Dilution Factor:
 Number TICs found:
   CONCENTRATION UNITS:
   (ug/L or ug/Kg)	
CAS NUMBER
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.
20.
21.
22.
23.
24.
25.
26.
27.
28.
29.
30.

COMPOUND NAME































RT































EST. CONC.





























•

Q


















"












                                FORM I SV-TIC
                                 3/90

-------
                              2A
          WATER VOLATILE SYSTEM MONITORING  COMPOUND  RECOVERY
 Lab Name:

 Lab Code:
Case No.:
Contract:

 SAS No.:
SDG No.
| EPA
| SAMPLE NO.
| ============
oil
02|
03|
04|
051
06|
07|
08|
09|
10|
HI
12|
131
14|
151
16|
17|
18|
191
20|
211
22|
23|
24|
25|
26|
27|
28|
29]
30|
SMC1
(TOL) |





























SMC2
(BFB) S





























SMC3
(DCE)#





























OTHER





























TOT
OUT





























                                                   QC LIMITS
               SMC1  (TOL) = Toluene-d8              (88-110)
               SMC2  (BFB) = Bromofluorobenzene      (86-115)
               SMC3  (DCE) = l,2-Dichloroethane-d4   (76-114)

               # Column to be used to flag  recovery  values

               * Values outside of contract required QC  limits

               D System Monitoring Compound diluted  out
page 	 of 	
                                FORM II VOA-1
                                                   3/90

-------
                              2B
          SOIL VOLATILE SYSTEM MONITORING COMPOUND RECOVERY
 Lab Name:

 Lab Code:
Case No.:
Contract:

 SAS No.:
SDG No.
 Level:(low/med)
| EPA
| SAMPLE NO.
oil
021
03|
04|
051
06|
07|
08|
09!
101
111
121
131
141
151
16|
17|
181
19|
20|
211
22|
23|
24|
251
26|
27|
28|
291
30|
SMC1
(TOL) S





























SMC2
(BFB)#
======





























SMC3
(DCE)#





























OTHER
'





























TOT|
OUT]
==>=





























                                                    QC LIMITS
               SMC1  (TOL)  =  Toluene-d8             (84-138)
               SMC2  (BFB)  =  Bromofluorobenzene     (59-113)
               SMC3  (DCE)  =  l,2-Dichloroethane-d4  (70-121)

               # Column  to be  used to flag recovery values

               * Values  outside  of contract required QC limits

               D System  Monitoring Compound diluted out
page 	 of 	
                                FORM  II  VOA-2
                                                   3/90

-------
                              2C
              WATER SEMIVOLATILE SURROGATE RECOVERY
 Lab Name:

 Lab Code:
Case No.:
Contract:

 SAS No.:
SDG No.
| EPA
| SAMPLE NO.
oil
02|
03|
04 |
05|
06|
07|
08|
09|
101
HI
12|
13|
14|
15|
16|
17|
18|
191
20|
211
22|
23|
24|
25|
26|
27|
28|
29|
30|
SI
(NBZ)S






























S2
(FBP)#






























S3
(TPH) *






























S4
(PHL) S






























S5
(2FP)#






























S6
(TBP) #






























S7
(2CP)#






























S8
(DCB) #






























TOT
OUT






























page 	 of 	
                                                   QC LIMITS
                 SI  (NBZ)  = Nitrobenzene-d5        (35-114)
                 S2  (FBP)  = 2-Fluorobiphenyl       (43-116)
                 S3  (TPH)  = Terphenyl-dl4          (33-141)
                 S4  (PHL)  = Phenol-d5              (10-110)
                 S5  (2FP)  = 2-Fluorophenol         (21-110)
                 S6  (TBP)  = 2,4,6-Tribroraophenol   (10-123)
                 S7  (2CP)  = 2-Chlorophenol-d4      (33--110)
                 S8  (DCB)  = l,2-Dichlorobenzene-d4 (16-110)
                                        (advisory)
                                        (advisory)
                 rt  Column to be used to flag recovery values
                 *  Values outside of contract required QC limits
                D Surrogate  diluted out
                                 FORM II  SV-1
                                                   3/90

-------
                              2D
               SOIL SEMIVOLATILE SURROGATE RECOVERY
 Lab  Name:

 Lab  Code:
Case No.:
Contract:

 SAS No.:
SDG No.:
 Level:(low/med)
| EPA
| SAMPLE NO.
oil
02|
03|
04 |
05|
06 |
07|
08|
09|
10|
HI
12|
13|
14|
15|
16|
17|
18|
191
301
211
2-2|
23|
24|
25|
26|
27|
28|
29|
30|
SI
(NBZ)#






























S2
(FBP)#















'














S3
(TPH) #






























S4
(PHL)#






























S5
(2FP)#






























S6
(TBP) #
1
1
1



























S7
(2CP)#


















.











S8
(DCB)#






























TOT
OUT






























                                                   QC LIMITS
                SI  (NBZ) = Nitrobenzene-d5         (23-120)
                S2  (FBP) = 2-Fluorobiphenyl        (30-115)
                S3  (TPH) = Terphenyl-dl4           (18-137)
                S4  (PHL) = Phenol-d5               (24-113)
                S5  (2FP) = 2-Fluorophenol          (25-121)
                S6  (TBP) = 2,4,6-Tribromophenol    (19-122)
                S7  (2CP) = 2-Chlorophenol-d4       (20-130)
                S8  (DCB) = l,2-Dichlorobenzene-d4  (20-130)
                                        (advisory)
                                        (advisory)
                # Column to be used  to  flag  recovery  values
                * Values outside  of  contract required QC limits
                D Surrogate diluted  out
page 	 of 	
                                 FORM  II SV-2
                                                                        3/90

-------
                             2E
                WATER PESTICIDE SURROGATE RECOVERY
Lab Name:

Lab Code:
GC Column(1):
   Contract:

    SAS  Mo.:
SDG No.:
(mm)   GC  Column(2):
         ID:
(mm)
| EPA
(SAMPLE NO.
|
oil
02 |
03|
04|
05|
06|
07|
08|
09 |
10|
HI
12|
13|
14|
15|
16|
17|
18|
19|
20|
211
22|
23|
24|
25|
26|
27|
28|
291
30|
TCX 1
%REC #






























TCX 2
%REC #






























DCB 1
%REC #






























DCB 2
%REC #






























OTHER
(1)

























t




OTHER
(2)






























TOT
OUT






























          TCX  =  Tetrachloro-m-xylene
          DCB  =  Decachlorobiphenyl
     ADVISORY
     QC LIMITS
     (60-150)
     (60-150)
          # Column  to  be  used to flag recovery values
          * Values  outside  of QC limits
          D Surrogate  diluted out
       of
                                FORM II  PEST-1
                                    3/90

-------
                              2F
                 SOIL PESTICIDE SURROGATE RECOVERY
 Lab Name:

 Lab Code:
 GC Column(1):
                              Contract:

                               SAS  No.:
SDG No.:
                           (mm)   GC Column(2):
         ID:
(mm)
| EPA
| SAMPLE NO.
1 ' '
oil
02|
03|
04|
05|
06|
07|
08)
09 |
10|
HI
12|
13|
14|
15|
16|
17|
18|
191
20|
211
22|
23|
24|
25|
26|
27|
28|
29|
30|
TCX 1
%REC g






























TCX 2
%REC #






























DCB 1
%REC #






























DCB 2
%REC #






























OTHER
(1)






























OTHER
(2)






























TOT]
OUT]






























           TCX = Tetrachloro-m-xylene
           DCB = Decachlorobiphenyl
                                ADVISORY
                                QC LIMITS
                                 (60-150)
                                 (60-150)
           $ Column to be used to  flag recovery values
           * Values outside of QC  limits
           D Surrogate diluted out
page
of
                                FORM II PE5T-2
                                                               3/90

-------
                              3A
         WATER VOLATILE  MATRIX SPIKE/MATRIX SPIKE DUPLICATE RECOVERY
 Lai Name:

 Lab Code:
Case No.:
Contract:

 SAS No.:
SDG No.:
Matrix Spike - EPA  Sample  No.:
COMPOUND
1 , 1-Dichloroethene
Trichloroethene
3enzene
Toluene
Chlorobenzene

SPIKE
ADDED
(ug/L)






| SAMPLE
| CONCENTRATION
1 (ug/L)
1 ~~
I
1
1
1
1
1
MS
CONCENTRATION
(ug/L)






MS
%
REC #






QC.
LIMITS
REC.
61-145
71-120
76-127
76-125
75-130

COMPOUND
1, 1-Dichloroethene
Trichloroethehe
Benzene
Toluene
Chlorobenzene
SPIKE
ADDED
(ug/L)





1
MSD
CONCENTRATION
(ug/L)






MSD | |
% | % | QC L]
REC #| RPD #| RPD
1 1 14
1 1 14
1 1 11
1 1 13
1 1 13
1 1
EMITS
REC.
61-145
71-120
76-127
76-125
75-130
r Column to be used to flag recovery  and  RPD  values  with an asterisk

* Values outside of QC limits

FJ'D:	 out of 	 outside  limits
Spike Recovery:	 out of 	 outside limits


COMMENTS:
                                 FORM  III  VOA-1
                                                   3/90

-------
                              3B
          SOIL VOLATILE  MATRIX SPIKE/MATRIX  SPIKE  DUPLICATE  RECOVERY
 Lab Name:

 Lab Code:
Case No.:
Contract:

 SAS No.:
SDG No.:
Matrix Spike - EPA  Sample  No.:
                          Level:(low/med)
COMPOUND
1, 1-Dichloroethene
Trichloroethene
Benzene
Toluene
Chlorobenzene

| SPIKE
| ADDED
1 (ug/Kg)
1
1
1
1
1
1
SAMPLE
CONCENTRATION
(ug/Kg)






MS | MS
CONCENTRATION | %
(ug/Kg) ! REC #
1
1
1
1
1
1
QC.
LIMITS
REC.
59-172
62-137
66-142
59-139
60-133

COMPOUND
1, 1-Dichloroethene
Trichloroethene
Benzene
Toluene
Chlorobenzene

SPIKE
ADDED
(ug/Kg)






MSD
CONCENTRATION
(ug/Kg)






MSD
%
REC #






%
RPD #
	





QC LJ
RPD
22
24
21
21
21

CMITS
REC. ,
59-172
62-137
66-142
59-139
60-133

# Column to be used to flag recovery and RPD values with an asterisk

* Values outside of QC limits

RPD:	 out of 	 outside limits
Spike Recovery:	 out of 	 outside limits


COMMENTS:
                                FORM III VOA-2
                                                   3/90

-------
                              3C
          WATER SEMIVOLATILE MATRIX  SPIKE/MATRIX SPIKE  DUPLICATE  RECOVERY
 Lab Name:

 Lab Code:
Case No.:
Contract:

 SAS No.:
SDG No.:
 Matrix Spike - EPA Sample No.:
COMPOUND
Phenol
2-Chlorophenol
1,4-Dichlorobenzene
N-Nitroso-di-n-prop. (1)
1,2, 4-Trichlorobenzene
4-Chloro-3-methylphenol
Acenaphthene
4-Nitrophenol
2 , 4-Dinitrotoluene
Pentachlorophenol
Pyrene

SPIKE
ADDED
(ug/L)












SAMPLE '
CONCENTRATION
(ug/L)












MS
CONCENTRATION
(ug/L)












MS
%
REC #












QC.
LIMITS
REC.
12-110
27-123
36- 97
41-116
39- 98
23- 97
46-118
10- 80
24- 96
9-103
26-127

COMPOUND
Phenol
2-Chlorophenol
1, 4-Dichlorobenzene
N-Nitroso-di-n-prop. (1)
1,2, 4-Trichlorobenzene
4-Chloro-3-methylphenol
Acenaphthene
4 -Ni trophenol
2 , 4-Dinitrotoluene
Pentachloropheno 1
Pyrene

SPIKE
ADDED
(ug/L)












MSD
CONCENTRATION
(ug/L)












MSD
%
REC #
	











%
RPD #












QC L:
RPD
42
40
28
38
28
42
31
50
38
50
31
CMITS
REC.
12-110
27-123
36- 97
41-116
39- 98
23- 97
46-118
10- 80
24- 96
9-103
26-127
 (1) N-Nitroso-di-n-propylamine

# Column to be used to flag recovery and RPD values with an asterisk
* Values outside of QC limits

RPD:	 out of 	 outside limits
Spike Recovery:	 out of 	 outside limits

COMMENTS:
                                FORM III SV-1
                                                   3/90

-------
                              3D
          SOIL SEMIVOLATILE MATRIX SPIKE/MATRIX SPIKE DUPLICATE RECOVERY
 Lab Name:

 Lab Code:
Case No.:
Contract:

 SAS No.:
SDG No.:
 Matrix Spike - EPA Sample No.:
                          Level:(low/med)
COMPOUND
Phenol
2-Chlorophenol
1 , 4-Dichlorobenzene
N-Nitroso-di-n-prop. (l)
1,2, 4-Trichlorobenzene
4-Chloro-3-methylphenol
Acenaphthene
4-Nitrophenol
2 , 4-Dinitrotoluene
Pentachlorophenol
Pyrene

SPIKE
ADDED
(ug/Kg)












SAMPLE '
CONCENTRATION
(ug/Kg)












MS
CONCENTRATION
(ug/Kg)












MS
%
REC #












QC.
LIMITS
REC.
26- 90
25-102
28-104
41-126
38-107
26-103
31-137
11-114
28- 89
17-109
35-142

COMPOUND
Phenol
2-Chlorophenol
1, 4-Dichlorobenzene
N-Nitroso-di-n-prop. (1)
1,2, 4-Trichlorobenzene
4-Chloro-3-methylphenol
Acenaphthene
4-Nitrophenol
2 , 4-Dinitrotoluene
Pentachlorophenol
Pyrene
.
SPIKE
ADDED
(ug/Kg)












MSD
CONCENTRATION
(ug/Kg)












MSD
%
REC #












%
RPD #












QC L]
RPD
35
50
27
38
23
33
19
50
47
47
36

EMITS
REC.
26- 90
25-102
28-104
41-126
38-107
26-103
31-137
11-114
28- 89
17-109
35-142

  (1) N-Nitroso-di-n-propylamine

# Column to be used to  flag  recovery  and RPD values  with  an  asterisk
* Values outside of QC  limits

RPD:	 out of 	 outside  limits
Spike Recovery:	 out of 	 outside limits

COMMENTS:
                                FORM III SV-2
                                                   3/90

-------
                              3E
          WATER PESTICIDE  MATRIX  SPIKE/MATRIX  SPIKE  DUPLICATE  RECOVERY
 Lab Name:

 Lab Code:
Case No.:
Contract:

 SAS No.:
                                    SDG No.:
 Matrix  Spike  - EPA Sample  No.:
COMPOUND
gamma-BHC (Lindane)
Heptachlor
Aldrin
Dieldrin
Endrin
4,4' -DDT

SPIKE
ADDED
(ug/L)
	






SAMPLE
CONCENTRATION
(ug/L)







MS
CONCENTRATION
(ug/L)







MS
%
REC #







QC. |
LIMITS
REC.
56-123
40-131
40-120
52-126
56-121
38-127

COMPOUND
gamma-BHC (Lindane)
Heptachlor
Aldrin
Dieldrin
Endrin
4,4' -DDT

SPIKE
ADDED
(ug/L)







MSD | MSD
CONCENTRATION | %
(ug/L) | REC #
1
1
1
1
1
1
1
%
RPD #







QC LIMITS
RPD | REC .
15 |56-123
20 (40-131
22 |40-120
18 |52-126
21 156-121
27 |38-127
1
I
# Column to be used to flag recovery and RPD values with  an  asterisk

* Values outside of QC limits
RPD:	 out of 	 outside limits
Spike Recovery:	 out of 	 outside limits
COMMENTS:
                                FORM III PEST-1
                                                   3/90

-------
                              3F
          SOIL PESTICIDE MATRIX SPIKE/MATRIX  SPIKE  DUPLICATE RECOVERY
 Lab Name:

 Lab Code:
Case No.:
Contract:

 SAS No.:
SDG No.:
 Matrix Spike - EPA Sample No.:
COMPOUND
ganuna-BHC (Lindane)
Heptachlor
Aldrin
Dieldrin
Endrin
4, 4 '-DDT

SPIKE
ADDED
(ug/Kg)







SAMPLE
CONCENTRATION
( ug/Kg)







MS
CONCENTPvATION
(ug/Kg)







MS
%
REC ft







QC.
LIMITS
REC.
46-127
35-130
34-132
31-134
42-139
23-134

COMPOUND
ganuna-BHC (Lindane)
Heptachlor
Aldrin
Dieldrin
Endrin
4,4' -DDT

| SPIKE
ADDED
(ug/Kg)







MSD
CONCENTRATION
(ug/Kg)







MSD
%
REC #







%
RPD #







QC L:
RPD
50
31
43
38
45
50

EMITS
REC.
46-127
35-130
34-132
31-134
42-139
23-134

I.
# Column to be used to  flag recovery and RPD values  with  an  asterisk

* Values outside of QC  limits
RPD:	 out of 	 outside limits
Spike Recovery:	 out of 	 outside  limits
COMMENTS:
                                FORM III PEST-2
                                                   3/90

-------
 Lab Name:

 Lab Code:
             4A
VOLATILE METHOD BLANK SUMMARY


                	   Contract:

                      SAS No.:
                                                              EPA SAMPLE NO.
Case No.:
          SDG No.:
 Lab File ID:

 Date Analyzed:

 GC Column:
                              Lab Sample ID:

                              Time Analyzed:
     ID:
           (mm)
Heated Purge:  (Y/N)
 Instrument ID:
       THIS  METHOD BLANK APPLIES TO THE FOLLOWING SAMPLES, MS AND MSD:
COMMENTS:
| EPA
| SAMPLE MO.
1
on
021
03|
04|
051
06|
07|
08|
09|
10|
HI
12|
131
14|
151
16|
17|
18|
19|
20|
211
22|
23|
24|
25|
26|
27|
28|
29|
30|
LAB
SAMPLE ID






























LAB
FILE ID






























TIME
ANALYZED






























page 	 of
                                 FORM IV VOA
                                                       3/90

-------
 Lab Name:

 Lab Code:
             4B
SEMIVOLATILE METHOD BLANK SUMMARY


                	   Contract:	

                      SAS No.:
                                                              EPA  SAMPLE NO.
Case No.:
SDG No.:
 Lab File ID:      	

 Instrument ID:

 Matrix:  (soil/water)

 Level:(low/med)
                             Lab Sample ID:

                             Date Extracted:

                             Date Analyzed:

                             Time Analyzed:
       THIS METHOD BLANK APPLIES TO THE FOLLOWING SAMPLES, MS AND  MSD:
| EPA
j SAMPLE NO.
on
021
031
04|
051
06|
07|
08|
09)
10|
111
121
131
141
151
161
17|
18| .
19|
20|
211
221
23|
241
25|
26|
27|
28|
291
30|
LAB
SAMPLE ID






























LAB
FILE ID






























DATE
ANALYZED






























COMMENTS:
page 	 of 	
                                 FORM IV SV
                                                       3/90

-------
 Lab Name:

 Lab Code:
             4C
PESTICIDE METHOD BLANK SUMMARY


                	   Contract:

                      SAS No.:
                                                              EPA SAMPLE NO.
Case No.:
SDG No.:
 Lab Sample ID:
                        Lab File ID:
 Matrix:(soil/water)
 Sulfur Cleanup:  (Y/N)  	

 Date Analyzed (1):  	

 Time Analyzed (1):  	

 Instrument ID (1):  	

 GC Column (1): 	
          ID:
                    Extraction:(SepF/Cont/Sonc)

                    Date Extracted:       	

                    Date Analyzed  (2):    	

                    Time Analyzed  (2):    	

                    Instrument ID  (2):    	

              (mm)   GC Column (2) :  	
          ID:
(mm)
          THIS  METHOD BLANK APPLIES TO THE FOLLOWING SAMPLES, MS AND MSD:

01
02
03
04
05
06
07
03
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
EPA
SAMPLE NO.





•

.?


















LAB
SAMPLE ID


























DATE
ANALYZED 1
	

























DATE |
ANALYZED 2


























COMMENTS:
page 	 of 	
                                 FORM IV PEST
                                                       3/90

-------
                              5A
         VOLATILE ORGANIC INSTRUMENT PERFORMANCE CHECK
                     BROMOFLUOROBENZENE (BFB)
 Lab Name:

 Lab Code:
Case No.:
      Contract:

       SAS No.:
             SDG No.:
 Lab File ID:

 Instrument ID:

 GC Column:
 ID:
(mm)
BFB Injection Date:

BFB Injection Time:

Heated Purge: (Y/N)
m/e
50
75
95
96
173
174
175
176
177

ION ABUNDANCE CRITERIA
8.0 - 40.0% of mass 95
30.0 - 66.0% of mass 95
Base peak, 100% relative abundance
5.0 - 9.0% of mass 95
Less than 2.0% of mass 174
50.0 - 120.0% of mass 95
4.0 - 9.0 % of mass 174
93.0 - 101.0% of mass 174
5.0 - 9.0% of mass 176

% RELATIVE
ABUNDANCE




( )1

( )1
( U
( )2

         1-Value is % mass 174
                          2-Value is % mass  176
 THIS CHECK APPLIES TO THE FOLLOWING SAMPLES, MS, MSD, BLANKS,  AND STANDARDS:
| EPA
| SAMPLE NO.
Oil
02 |
03|
04|
05|
06|
07|
08|
09|
10|
HI
12|
13|
14|
15|
16|
17|
18|
19|
20|
21!
22|
LAB
SAMPLE ID






















LAB
FILE ID






















DATE
ANALYZED






















TIME
ANALYZED






















page 	 of 	
                                 FORM V VGA
                                                                     3/90

-------
                              5B
         SEMIVOLATILE ORGANIC INSTRUMENT PERFORMANCE CHECK
               DECAFLUOROTRIPHENYLPHOSPHINE (DFTPP)
 Lab Name:
                 Contract:
 Lab Code:
Case No.:
SAS No.:
SDG No.:
 Lab File ID:
 Instrument ID:
                     DFTPP Injection Date:

                     DFTPP Injection Time:

1
1 n/e
I' 51
| 68
| 69
[ 70
| 127
| 197
| 198
f 199
[ 275
| 365
| 441
|: 442
| 443
1
- -
ION ABUNDANCE CRITERIA
30.0 - 80.0% of mass 198
Less than 2.0% of mass 69
Mass 69 relative abundance
Less than 2.0% of mass 69
25.0 - 75.0% of mass 198
Less than 1.0% of mass 198
Base Peak, 100% relative abundance
5.0 to 9.0% of mass 198
10.0 - 30.0% of mass 198
Greater than 0.75% of mass 198
Present, but less than mass 443
40.0 - 110.0% of mass 198
15.0 - 24.0% of mass 442


% RELATIVE
ABUNDANCE

( ) 1

( )1








( )2

         1-Value is % mass 69
                          2-Value is % mass 442
 THIS  CHECK APPLIES TO THE FOLLOWING SAMPLES, MS, MSD, BLANKS, AND  STANDARDS:
| EPA
| SAMPLE NO.
oil
02|
03|
04|
05|
06|
07|
08|
09|
10|
HI
12|
13|
14|
15|
16|
17|
18|
19|
20)
211
22|
LAB
SAMPLE ID






















LAB
FILE ID






















DATE
ANALYZED






















TIME
ANALYZED





















-
page 	 of 	
                                 FORM V SV
                                                                    3/90

-------
                               6A
                VOLATILE ORGANICS INITIAL CALIBRATION DATA
 Lab Name:
                 Contract:
 Lab Code:
Case No.:
SAS No.:
                                    SDG No.:
 Instrument ID:
 Heated Purge:  (Y/N)  	

 GC Column:            ID:
       Calibration Date(s):

       Calibration Times:

      	(mm)
[LAB FILE ID: RRF10 = RRF20 =
[RRF50 = RRF100= RRF200=
1
1 I
1 COMPOUND IRRFIO
1 Chloromethane |
| Bromomethane *
| Vinyl Chloride *
| Chloroethane |
IMethylene Chloride |
| Acetone |
(Carbon Disulfide I
| 1 , 1-Dichloroethene *
| 1 , 1-Dichloroethane *
| 1,2-Dichloroethene (total)
| Chloroform *
| 1, 2-Dichloroethane *
| 2-Butanone |
| 1 , 1 , 1-Trichloroethane *
| Carbon Tetrachloride *
| Bromodichloromethane *
j 1, 2-Dichloropropane |
|cis-l,3-Dichloropropene *
| Trichloroethene *
| Dibromochloromethane *
| 1, 1,2-Trichloroe thane *
| Benzene *
| trans-l,3-Dichloropropene *
| Bromof orm *
| 4-Methyl-2-Pentanone
| 2-Hexanone
| Tetrachloroethene *
| 1, 1,2, 2-Tetrachloroe thane *
| Toluene *
(Chlorobenzene *
| Ethylbenzene *
| Styrene *
IXylene (tota,l) *
|Toluene-d8 j
| Bromof luorobenzene *
| 1, 2-Dichloroethane-d4
1
RRF20





































RRF50











•

























RRF100





































RRF200
	 	 —




































•
RRF






































%
RSD

*
*
	


*
*
1
*
*
i

*
*
1
	 *
*
*
*
*
*
*
1

*
*
*
*
*
*
*

*


* Compounds with required minimum RRF and maximum  %RSD values.
  All other compounds must meet a minimum RRF of 0.010.
                                FORM VI VOA
                                               3/90

-------
                              6B
           SEMIVOLATILE ORGANICS  INITIAL  CALIBRATION  DATA
 Lab Name:

 Lab Code:
Case No.:
Contract:

 SAS No.:
SDG No.
 Instrument  ID:
       Calibration Date(s):

       Calibration Times:
ILAB FILE ID: RRFZO = RRFSO =
IRRF80 = RRF120= RRF160=
1
1 1
| COMPOUND |RRF20
| Phenol *
j bis (2-Chloroethyl) ether *
| 2-Chlorophenol *
j 1,3-Dichlorobenzene *
| 1 , 4-Dichlorobenzene *
1 1 , 2-Dichlorobenzene *
| 2-Methylphenol *
|2,2'-oxybis(l-Chloropropane) |
| 4-Methylphenol *
|N-Nitroso-di-n-propylamine *
| Hexachloroethane *
j Nitrobenzene *
| Isophorone *
| 2-Nitrophenol *
j 2 , 4-Dimethylphenol *
|bis(2-Chloroethoxy)methane *
| 2,4-Dichlorophenol *
(1,2,4-Trichlorobenzene *
| Naphthalene *
| 4-Chloroaniline " |
| Hexachlorobutadiene |
|4-Chloro-3-methylphenol *
| 2-Methylnaphthalene *
|Hexachlorocyclopentadiene |
| 2,4, 6-Trichlorophenol *
|2,4,5-Trichlorophenol *
|2-Chloronaphthalene *
|2-Nitroaniline |
| Dimethylphthalate |
| Acenaphthylene *
|2,6-Dinitrotoluene *
| 3-Nitroaniline |
| Acenaphthene *
| 2,4-Dinitrophenol |
|4-Nitrophenol j-
| Dibenzof uran *
| 2 , 4-Dinitrotoluene *
1 1
RRFSO






































RRFSO






































RRF120






































RRF160







































RRF







































% 1
RSD |
	 1
1
*
*
*
it
*
*
*
1
*
*
*
*
*
it
*
*
*
*
*
1
1
*
*
1
*
*
*
1
1
*
*
1
*
1
1
*
it
1
* Compounds with required minimum RRF and maximum %RSD values.
  All other compounds must meet a minimum RRF of 0.010.
                                FORM VI SV-1
                                               3/90

-------
                             6C
          SEMIVOLATILE ORGANICS INITIAL CALIBRATION DATA
Lab Name:

Lab Code:
Case No.:
Contract:

 SAS No.:
SDG No.
Instrument ID:
       Calibration Date(s):

       Calibration Times:
LAB FILE ID: RRF20 =
RRF80 = RRF120=


COMPOUND
Diethylphthalate
4-Chlorophenyl-phenylether <
Fluorene <
4-Nitroaniline
4 , 6-Dinitro-2-methylphenol
N-Ni trosodipheny lamine ( l )
4-Bromophenyl-phenylether '
Hexachlorobenzene '
Pentachlorophenol *
Phenanthrene 1
Anthracene '
Carbazole
Di-n-butylphthalate
Fluoranthene *
Pyrene *
Butylbenzylphthalate
3,3' -Dichlorobenzidine
Ben z o ( a ) anthracene *
Chrysene *
bis (2-Ethylhexyl) phthalate
Di-n-octylphthalate
Benzo(b) fluoranthene *
Ben z o ( k ) f luor anthene *
Benzo(a)pyrene *
Indeno(l,2,3-cd)pyrene *
Dibenz (a ,h) anthracene *
Benzo(g,h,i)perylene *

Nitrobenzene-d5
2-Fluorobiphenyl *
Terphenyl-dl4 *
Phenol-d5 *
2-Fluorophenol <
2,4, 6-Tribromophenol
2-Chlorophenol-d4 J
l,2-Dichlorobenzene-d4 *

a
3


RRF20

k
k



k
k
k
k
k


k
It


k
k


k
k
k
k
k
k


k
k
k
k

k
k





RRF50





































RRF50
RRF16(


RRF80





































s
)=


RRF120









































RRF160
































1








RRF








































%
RSD

*
4



A
*
*
*
4
i
(

A


4
*


4
4
4
4
*
y

i
1
i
i
•i

1
•*

(1)  Cannot be separated from Diphenylamine
* Compounds with required minimum RRF and maximum %RSD values.
  All other compounds must meet a minimum RRF of 0.010.
                                FORM VI SV-2
                                               3/90

-------
                             6D
        PESTICIDE INITIAL CALIBRATION OF SINGLE COMPONENT ANALYTES
 ab Name:

Lab Code:
Instrument ID:

GC Column:
Case No.:
Contract:

 SAS No.:
 	  Level (x low):  low 	

  ID: 	(mm)   Date(s)  Analyzed:
	  SDG No.: _

mid 	 high
COMPOUND
alpha-BHC
beta-BHC
delta-BHC
gamma-BHC (Lindane)
Heptachlor
Aldrin
Heptachlor epoxide
Endosulfan I
Dieldrin
4,4' -DDE
Endrin
Endosulfan II
4, 4 '-ODD
Endosulfan sulfate
4,4' -DDT
Methoxychlor
Endrin ketone
Endrin aldehyde
alpha-Chlordane
gamma-Chlordane
Tetrachloro-m-xylene
Decachlorobiphenyl

RT 0
LOW























F STAND;
MID























\RDS
HIGH























MEAN
RT























RT w:
FROM























LNDOW
TO




















	


     * Surrogate retention times are measured from Standard Mix A analyses.

     Retention  time windows are +0.05 minutes for all compounds that elute
     before  Heptachlor epoxide, +0.07 minutes for all other compounds,
     except  +0.10 minutes for Decachlorobiphenyl.
                                FORM VI PEST-1
                                              3/90

-------
                              6E
        PESTICIDE  INITIAL CALIBRATION  OF  SINGLE  COMPONENT ANALYTES
 Lab Name:

 Lab Code:
Case No.:
           Contract:

            SAS  No.:
 Instrument  ID:

 GC Column:
  ID:
 Level  (x  low):  low

	(mm)    Date(s)  Analyzed:
COMPOUND
alpha-BHC
beta-BHC
delta-BHC
gamma-BHC (Lindane)
Heptachlor
Aldrin
Heptachlor epoxide
Endosulfan I
Dieldrin
4,4 '-DDE
Endrin
Endosulfan II
4, 4 '-ODD
Endosulfan sulfate
4,4' -DDT
Methoxychlor
Endrin ketone
Endrin aldehyde
alpha-Chlordane
gamma-Chlordane
Tetrachloro-m-xylene
Decachlorobiphenyl

LOW























CALIBRATK
MID






















DN FACTORS
HIGH






















1
MEAN




















	


%RSD























* Surrogate calibration factors are measured  from Standard  Mix A analyses.

;RSD must be less than or equal 20.0 %  for all compounds  except the
-.surrogates, where %RSD must be less than or equal to  30.0%.   Up to
>:wo target compounds, but not surrogates, may have  %RSD greater than
20.0% but less than or equal to 30.0%.
                                 FORM VI  PEST-2
                                                 3/90

-------
                        6F
 PESTICIDE INITIAL CALIBRATION OF MULTICOMPONENT ANALYTES
Lab Name :
Lab Code :
Instrument ID:
GC Column:
Contract:
Case No.: SAS No.: SDG No.:
Date(s) Analyzed:
ID: (mm)
COMPOUND
Toxaphene
Aroclor 1016
Aroclor 1221
.
Aroclor 1232
Aroclor 1242
Aroclor 1248
Aroclor 1254
Aroclor 1260
AMOUNT
(ng)
















PEAK
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5
RT









































RT w:
FROM
	








































[NDOW
TO









































CALIBRATION
FACTOR









































* Denotes required peaks
                           FORM VI  PEST-3
3/90

-------
                              GG
                 PESTICIDE ANALYTE  RESOLUTION  SUMMARY
Lab Name:

Lab Code:
Case No.:
   Contract:

    SAS No.:
                                    SDG No.:
GC Column  (1):
          01
          02
          03
          04
          05
          06
          07
          08
          09
: ID: (mm) ]
>. (Standard 1) : I
(1): -3

ANALYTE









Cnstrunu
jab Samj
rime An;

RT









snt ID (1) :
ale ID (1) :
alyzed (1):

RESOLUTION
(%)









GC Column (2):
      ID:
EPA Sample No. (Standard 2) :

Date Analyzed (2): 	
 (mm)  Instrument ID  (2):

	 Lab Sample ID  (2):

      Time Analyzed  (2):

01
02
03
04
05
06
07
08
09
ANALYTE









RT









RESOLUTION
(%)









  Resolution of two adjacent peaks must be calculated as a percentage of the
  height  of the smaller peak, and must be greater th?.- ->r equal  to  60.0%.
                                FORM VI PEST-4
                                               3/90

-------
                             7A
            VOLATILE CONTINUING CALIBRATION CHECK
Lab Name :
Lab Code:
Instrument ID:

Case No. :
Contract:
SAS No- :
Calibration Date:

SDG No. :
Time:
Lab File ID:
Heated Purge: (Y/N) 	

GC Column:           ID:
 Init. Calib.  Date(s):

 Init. Calib.  Times:

	(mm)
1
| COMPOUND.
j Chlorome thane
| Bromomethane
| Vinyl Chloride
| Chloroethane
IMethylene Chloride
| Acetone
| Carbon Disulfide
| 1, 1-Dichloroethene
| 1, 1-Dichloroethane
1 1 , 2-Dichloroethene ( total )
| Chloroform
1 1, 2-Dichloroethane
| 2-Butanone
| 1 , 1 , 1-Trichloroethane
| Carbon Tetrachloride
| Bromodichloromethane
| 1 , 2-Dichloropropane
| cis-1, 3-Dichloropropene
| Trichloroethene
| Dibromochloromethane
| 1 , 1 , 2-Trichloroethane
| Benzene
| trans-l, 3-Dichloropropene
| Bromof orm
| 4-Methyl-2-Pentanone
| 2-Hexanone
| Tetrachloroethene
| 1 , 1 , 2 , 2-Tetrachloroethane
(Toluene
| Chlorobenzene
| Ethylbenzene
| Styrene
IXylene (total)
|Toluene-d3
| Bromof luorobenzene
| 1, 2-Dichloroethane-d4
1

RRF






































RRF50





































MIN
RRF %D
	 	
0.100
0.100|
1
1


0.100
0.200

0.200
0.100
1
0.100
0.100
0.200

0.200
0. 300|
0.100
0.100
0.500
0.100
0. 100


0.200
0.500
0.400
0.500|
0.100
0.300 •
0.300

0.200


MAX)
%D |
i
	 1
1
25.0|
25.0|
1
1
1
1
25.0|
25.0|
1
25.0|
25.0|
1
25.0|
25.0|
25.0|
1
25.0|
25.0|
25.0|
25.0|
25.0|.
25.0|
25.0|
1
1
25.0)
25.0|
25.0|
25.0)
25.0|
25.0|
25.0|
1
1
25.0|
1
• 1
     All  other compounds must meet a minimum RRF of 0.010.

                                FORM VII VGA
                                        3/90

-------
                             7B
          SEMIVOLATILE CONTINUING CALIBRATION  CHECK
,ab Name:
,ab Code:
instrument ID:

Case No. :
Contract:
SAS No . :
Calibration Date:

SDG No. :
Time:
,ab  File  ID:
Init. Calib. Date(s):

Init. Calib. Times:

COMPOUND
Phenol
bis (2-Chloroethyl) ether
2-Chlorophenol
1 , 3 -Dichlorobenzene
1 , 4-Dichlorobenzene
1 , 2 -Dichlorobenzene
2-Methylphenol
2,2' -oxybis ( 1-Chloropropane)
4 -Methy Iphenol
N-Nitroso-di-n-propylamine
Hexachloroethane
Nitrobenzene
Isophorone
2-Nitrophenol
2 , 4-Dimethylphenol
bis ( 2-Chloroethoxy ) methane
2 , 4-Dichlorophenol
1,2,4 -Tr ichlorobenzene
Naphthalene
4-Chloroaniline
Hexachlorobutadiene
4 -Chloro-3 -methylphenol
2-Methylnaphthalene
Hexachlorocyclopentadiene
2,4, 6-Trichlorophenol
2,4, 5-Trichlorophenol
2-Chloronaphthalene
2-Nitroaniline
D ime thy Iphtha late
Acenaphthylene
2 , 6-Dinitrotoluene
3-Nitroaniline
Acenaphthene
2 , 4 -Dinitrophenol
4-Nitrophenol
Dibenzofuran
2 , 4 -Dinitrotoluene


RRF







































RRF50






































MIN
RRF %D
0.800
0.700
0.800
0.600
0.500
0.400
0.700

0.600
0.500
0.300
0.200
0.400
0.1001
0.200|
0.3001
0.200
0.200
0.700


0.200|
0.4001

0.200
0.200|
0.800


1.300
0.200
1
0.800|
1

0.800
0.200

MAX
%D
25.0
25.0
25.0
25.0
25.0
25.0
25.0

25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0


25.0
25.0

25.0
25.0
25.0


25.0
25.0

25.0


25.0
25.0

     A LI other compounds  must meet a minimum RRF of 0.010.
                                FORM VII SV-1
                                       3/90

-------
                             7C
          SEMIVOLATILE CONTINUING CALIBRATION CHECK
>b Name:
Contract:
l,ab Code: • Case No.: SAS No.: SDG No. :
Instrument ID:
Calibration Date: Time:
Lab File ID:
Init. Calib. Date(s):

Init. Calib. Times:
1
| COMPOUND
| Diethylphthalate
| 4-Chlorophenyl-phenylether
| Fluorene
| 4-Nitroaniline
| 4 , 6-Dinitro-2-methylphenol
| N-Nitrosodipheny lamina (1)
[ 4-Bromophenyl-phenylether
| Hexachlorobenzene
| Pentachlorophenol
| Phenanthrene
| Anthracene
| Carbazole
| Di-n-butylphthalate
| Fluoranthene
| Pyrene
| Butylbenzylphthalate
| 3 , 3 ' -Dichlorobenzidine
| Benzo (a) anthracene
| Chrysene
| bis (2-Ethylhexyl) phthalate
| Di-n-octylphthalate
| Benzo (b) fluoranthene
| Benzo (k) fluoranthene
| Benzo (a) pyrene
| Indeno (1,2,3 -cd) pyrene
| Dibenz (a , h) anthracene
| Benzo (g,h, i) perylene
| 	 :
| Nitrobenzene-d5
2-Fluorobiphenyl
Terphenyl-dl4
Phenol-d5
2-Fluorophenol
2,4, 6-Tribromophenol
2-Chlorophenol-d4
1 , 2-Dichlorobenzene-d4


RRF





































RRF50




































MIN |
RRF | %D
	 1 	
1
1
0.400|
0.900|
1
1
1
0.100|
0.1001
0.050|
0.700|
0.700|
1
1
0.600|
0.600|
1
1
0.800|
0.700|
1
1
0.700|
0.700|
0.700|
0.500|
0.400|
0.500|
= 	 1 	
0.200|
0.700]
0.500|
0.800|
0.600|
1
0.800|
0.400|
1
MAX
%D

25.0
25.0



25.0
25.0
25.0
25.0
25.0


25.0
25.0


25.0
25.0


25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0

25.0
25.0

    1) Cannot be  separated from Diphenylamine
     All other compounds must meet a minimum RRF of 0.010.
                                FORM VII SV-2
                                       3/90

-------
                              7D
               PESTICIDE CALIBRATION  VERIFICATION  SUMMARY
 Lab Name:_

 Lab Code:

 GC Column:
Contract:

 SAS No.:
Case No.:  	  SAS No.:  	  SDG No.:

  ID: 	(mm)  Init. Calib.  Date(s):
 EPA  Sample  No.(PIDLK):

 Lab  Sample  ID  (PIBLK):

 EPA  Sample  No.(PEM):  _

 Lab  Sample  ID  (PEM):	
         Date Analyzed

         Time Analyzed

         Date Analyzed

         Time Analyzed
PEM
COMPOUND
alpha-BHC
beta-BHC
gamma-BHC (Lindane)
Endrin
-4,4' -DDT
Methoxychlor

RT







RT w:
FROM







[NDOW
TO







CALC
AMOUNT
(ng)







NOM
AMOUNT
(ng)
~






RPD




	

4,4'-DDT % breakdown  (1):

Combined % breakdown  (1):
    Endrin %  breakdown (1):
QC LIMITS:

     RPD of amounts in PEM must be less than or equal to 25.0%

     4,4'-DDT breakdown must be less than or equal to 20.0%

     Endrin breakdown must be less than or equal to 20.0%

     Combined breakdown must be less than or equal to 30.0%
                                FORM VII PEST-1
                              3/90

-------
                             7E
              PESTICIDE CALIBRATION VERIFICATION SUMMARY
Lab Name:_

Lab Code:

GC Column:

EPA Sample No.(PIBLK):

Lab Sample ID  (PIBLK):

EPA Sample No.(INDA):

Lab Sample ID  (INDA):
                       Contract:

                        SAS No.:
SDG No.:
                  (mm) Init. Calib. Date(s):
                                Date Analyzed  :

                                Time Analyzed  :

                                Date Analyzed  :

                                Time Analyzed  :
INDIVIDUAL MIX A
COMPOUND
alpha-BHC
qamma-BHC (Lindane)
Heptachlor
Endosulfan I
Dieldrin
Endrin
4, 4 '-ODD
4,4' -DDT
Methoxychlor
Tetrachloro-m-xylene
Decachlorobiphenyl

RT












RT W]
FROM












ENDOW
TO












CALC
AMOUNT
(ng)












NOM
AMOUNT
(ng)












RPD










*

EPA Sample No.(INDB)

Lab Sample ID (INDB)
                                Date Analyzed  :

                                Time Analyzed  :
INDIVIDUAL MIX B
COMPOUND
beta-BHC
delta-BHC
Aldrin
Heptachlor epoxide
4, 4 '-DDE
Endosulfan II
Endosulfan sulfate
Endrin ketone
Endrin aldehyde
alpha-Chlordane
gamma-Chlordane
Tetrachloro-m-xylene
Decachlorobiphenyl

RT














RT W]
FROM














ENDOW
TO














CALC
AMOUNT
(ng)














NOM
AMOUNT
(ng)













-
RPD














  QC LIMITS:
RPD of amounts in the Individual Mixes must be  less  than
or equal to 25.0%.
                                FORM VII PEST-2
                                                      3/90

-------
                              8A
               VOLATILE INTERNAL STANDARD AREA AND RT SUMMARY
 Lab Name:

 Lab Code:
Case No.:
Contract:

 SAS No.:
SDG No.:
 Lab File ID (Standard):

 Instrument ID: 	

 GC Column:           ID:
                            Date Analyzed:

                            Time Analyzed:
           (mm)
           Heated Purge:  (Y/N)






01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22

12 HOUR STD
UPPER LIMIT
LOWER LIMIT

EPA SAMPLE
NO.






















ISl(BCM)
AREA I



























RT f



























IS2(DFB)
AREA #



























RT *



























IS3(CBZ)
AREA #















-











RT S



























    IS1  (BCM) = Bromochloromethane
    IS2  (DFB) = 1,4-Difluorobenzene
    IS3  (CBZ) = Chlorobenzene-d5

    AREA UPPER LIMIT =  +100%  of internal  standard  area
    AREA LOWER LIMIT =  -  50%  of internal  standard  area
    RT UPPER LIMIT = +0.50 minutes of  internal  standard  RT
    RT LOWER LIMIT = -0.50 minutes of  internal  standard  RT

    » Column used to flag values  outside  QC  limits with  an  asterisk.
    * Values outside of QC limits.
page 	.of 	
                                 FORM VIII VOA
                                               3/90

-------
                              8B
               SEMIVOLATILE INTERNAL STANDARD AREA  AND  RT  SUMMARY
 Lab Name:

 Lab Code:
Case No.:
Contract:

 SAS No.:
SDG No.:
 Lab File ID (Standard):

 Instrument ID:
                            Date  Analyzed:

                            Time  Analyzed:





01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22

12 HOUR STD
UPPER LIMIT
LOWER LIMIT
EPA SAMPLE
NO.






















ISl(DCB)
AREA I


























RT #


























IS2(NPT)
AREA g


























RT #


























IS3(ANT)
AREA g


























RT #


























    IS1 (DCB) = l,4-Dichlorobenzene-d4
    IS2 (NPT) = Naphthalene-d8
    IS3 (ANT) = Acenaphthene-dlO

    AREA UPPER LIMIT = +100% of internal standard area
    AREA LOWER LIMIT = - 50% of internal standard area
    RT UPPER LIMIT = +0.50 minutes of internal standard RT
    RT LOWER LIMIT = -0.50 minutes of internal standard RT

    if Column used to flag internal standard area values with  an-asterisk.
    * Values outside of QC limits.
page 	 of 	
                                FORM VIII SV-1
                                               3/90

-------
                              8C
               SEMIVOLATILE INTERNAL STANDARD  AREA AND  RT  SUMMARY
 Lab Name:

 Lab Code:
Case No.:
 Lab File ID (Standard):

 Instrument ID:
Contract:

 SAS No.:
SDG No.
                            Date Analyzed:

                            Time Analyzed:
1
1
| 12 HOUR STD
| UPPER LIMIT
| LOWER LIMIT
| 	
| EPA SAMPLE
| NO.
oil
021
03|
04 |
05|
06|
07|
08|
09|
10|
HI
12|
13|
14|
15|
161
17|
18|
19|
20|
211
22|
IS4(PHN)
AREA £
1

1























RT s


























ISS(CRY)
AREA £


























RT S


























IS6(PRY)
AREA #


























RT #


























    IS4  (PHN) = Phenanthrene-dlO
    IS5  (CRY) = Chrysene-dl2
    IS6  (PRY) = Perylene-dl2

    AREA UPPER LIMIT = +100% of internal standard area
    AREA LOWER LIMIT = - 50% of internal standard area
    RT UPPER LIMIT = +0.50 minutes of internal standard RT
    RT LOWER LIMIT = -0.50 minutes of internal standard RT

    tf Column used to flag internal standard area values with an  asterisk.
    * Values outside of QC limits.

page 	 of 	
                                FORM VIII SV-2
                                               3/90

-------
 Lab Name:_

  ab Code:

 GC Column:
           8D
 PESTICIDE  ANALYTICAL SEQUENCE

	   Contract:	

 Case  No.: 	  SAS No.: 	  SDG  No.:

	  ID: 	(mm) Init. Calib. Date(s):	
 Instrument ID:
   THE ANALYTICAL SEQUENCE OF PERFORMANCE EVALUATION MIXTURES,  BLANKS,
                SAMPLES, AND STANDARDS IS GIVEN BELOW:
| MEAN SURROGATE RT FROM INITIAL CALIBRATION
| TCX: DCB:
1
| EPA
| SAMPLE NO.
i
1 	
02|
03|
04 |
05|
06|
07|
08|
09 |
10|
HI
12 |
13 |
14 |
15 |
16 |
17 |
18 |
19 |
20|
21 1
22|
23|
24|
25|
26|
27|
28|
29|
30|
311
32|
LAB
SAMPLE ID


















•













DATE TIME
ANALYZED ANALYZED
— __. 1
	 1 	
1
1
1
1
1
'



















































TCX
RT $
	































DCB
RT jf
































        TCX  =  Tetrachloro-m-xylene
        DCB  =  Decachlorobiphenyl
                     QC LIMITS
                  (+ 0.05 MINUTES)
                  (+ 0.10 MINUTES)
        = Column  used to flag retention time values with an asterisk.
        * Values  outside of QC Limits.
page
        of
             FORM VIII PEST
3/90

-------
                              9A
                  PESTICIDE  FLORISIL  CARTRIDGE CHECK
 ;,ab Name:

 Lab Code:
Case No.:
Contract:

 SAS No.:
                                    S DG  No. :
 Florisil Cartridge Lot Number:

 GC Column(l): 	 ID:
                      Date of Analysis:
              (mm)  GC Column(2):
                             ID:
                                      (mm)
COMPOUND
alpha-BHC
gamma-BHC (Lindane)
Heptachlor
Endosulfan I
Dieldrin
Endrin
4,4' -ODD
4,4' -DDT
Methoxychlor
Tetrachloro-m-xylene
Decachlorob ipheny 1

SPIKE
ADDED
(ng)












SPIKE
RECOVERED
(ng)












%
REC #












QC
LIMITS
80-120
80-120
80-120
80-120
80-120
80-120
80-120
80-120
80-120
80-120
80-120

      % Column to be used to flag recovery with an  asterisk.
      * Values outside of QC limits.

  THIS CARTRIDGE LOT APPLIES TO THE FOLLOWING SAMPLES,  BLANKS,  MS, AND MSD:
| EPA
| SAMPLE NO.
oil
02|
03|
04 |
05|
06|
07|
08|
09|
10|
HI
12|
13|
14 |
15|
16|
17 |
18|
19|
20|
21|
22|
23|
LAB
SAMPLE ID























DATE
ANALYZED 1























DATE
ANALYZED 2























:iage 	 of 	
                                  FORM IX PEST-1
                                                 3/90

-------
                              9B
                  PESTICIDE GPC CALIBRATION
 Lab Name:

 Lab Code:
 GPC Column:

 GC Column(1):
Case No.:
     ID:
   Contract:

    SAS  No.:
SDG No.:
    Calibration Date:

(mm)   GC Column(2):
         ID:
(mm)
COMPOUND
gamma-BHC (Lindane)
Heptachlor
Aldrin
Dieldrin
Endrin
4,4' -DDT

SPIKE
ADDED
(ng)







SPIKE
RECOVERED
(ng)







REC j?







QC. |
LIMITS |
REC. j
80-1101
80-1101
80-110|
80-1101
80-110J
80-110 1
1
           # Column to be used to flag recovery values with an asterisk
           * Values outside of QC limits

   THIS  GPC CALIBRATION APPLIES TO THE FOLLOWING SAMPLES, BLANKS, MS AND  MSD:
| EPA
| SAMPLE NO.

02|
03|
04 |
05|
06|
07 |
08 |
09|
10 |
HI
12|
13 |
14 |
15 |
16 |
17 |
18 |
191
20|
211
22|
23|
24|
25|
26|
LAB
SAMPLE ID


























DATE
ANALYZED 1


























DATE
ANALYZED 2
























•

page 	 of 	
                                 FORM IX PEST-2
                                               3/90

-------
                              10A
                   PESTICIDE IDENTIFICATION SUMMARY
                    FOR SINGLE COMPONENT ANALYTES
                                                              EPA SAMPLE NO.
 Lab Name:

 Lab Code:
Case No.:
   Contract:

    SAS No.:
SDG No.:
 Lab Sample ID :
 Instrument ID (1):

 GC Column(1): 	
     ID:
      Date(s)  Analyzed:

      Instrument ID (2):

(mm)   GC Column(2): 	
                                              ID:
                   (mm)
ANALYTE









COL | RT
1
2
1
1
2
1
2
1
2
1
2
1
2
1
2

















RT w:
FROM












.

ENDOW
TO














!
!
_

CONCENTRATION
















•
%D









page 	 of
                                 FORM X PEST-1
                                                3/90

-------
                              10B
                   PESTICIDE IDENTIFICATION SUMMARY
                     FOR MULTICOMPONENT ANALYTES
                                         EPA SAMPLE NO.
 Lab Name:

 Lab Code:
Case No.:
   Contract:

    SAS No.:
SDG No.:
 Lab Sample ID :
 Instrument ID (1) :

 GC Column(1):  	
     ID:
      Date(s)  Analyzed:

      Instrument ID (2):

(mm)   GC Column(2): 	
         ID:
(mm)
ANALYTE

COLUMN 1
COLUMN 2

COLUMN 1
COLUMN 2

COLUMN 1
COLUMN 2
1
PEAK | RT
1
2
3
4
5
1
2
3
4
5
1
2
3
4
5
1
3
4
5
1 '
2
3
4
5
1
2
3
4
5































RT W.
FROM




















	










ENDOW
TO





























CONCENTRATION





























1

MEAN
CONCENTRATION


i




%D




 At least 3 peaks are required  for  identification of multicomponent analytes

page 	 of 	

                                 FORM  X  PEST-2                       3/90

-------
SAMPLE LOG-IN SHEET
Lab Name: Pace of
Received By (Prim Name): Log-in Dale:
•

CMC Number:
Sample Delivery
Group No.:
SAS Number

REMARKS:
1. Custody Seal(s) Present/Absent*
Intact/Broken
2. Custody Seal Nou:

3. Qiaim-of -Custody Present/Absent*
Records
4. Traffic Reports or Present/Absent*
Packing Lia
5. Airbill Airbill/Slicker
Presenl/Absenl*
6. Airbill No.:

7. Sample Tags Present/Absent*
Sample Tag Listed/Not Ligi-d
•: Numbers on Chain-of-
Custody
1 8. Sample Condition: Intact/Broken*/
' Leaking
: 9. Does infoimatioa on
reports, and sample

1 0. Dale Received at Lab:

1 1. Time Received:

Sample Transfer


By: 	 	
On:

EPA
SAMPLE
9


























CORRESPONDING
SAMPLE
TAG
0

























ASSIGNED
LAB
tt


























REMARKS:
CONDITION
OF SAMPLE
SHIPMENT, ETQ.

























* Contact SMO and allaca record of resolution
Reviewed  By:  	
Date:  	
Logbook No.:  	
Logbook Page No:
                                                      FORMDC-1

-------
                        ORGANICS COMPLETE SDG FILE (CSF)  INVENTORY SHEET
LABORATORY NAME
CITY/STATE
CASE NO. SDG NO.
SDG NOS. TO FOLLOW
SAS NO.
CONTRACT NO.
SOW NO.

     All documents  delivered in  the  complete SDG  file must  be original  documents
     where possible.  (REFERENCE EXHIBIT B,  SECTION  II  and  SECTION III.)
                                                                PAGE NOs         CHECK
                                                               FROM   TO      LAB     EPA
1.   Inventory Sheet (Form DC-2)  (Do not number)
2.   SDG Case Narrative
3.   SDG Cover Sheet/Traffic Report
4.   Volatiles Data
     a.  QC Summary
         System Monitoring Compound Summary
           (Form II  VGA)
         Matrix Spike/Matrix Spike Duplicate Summary
           (Form III VOA)
         Method Blank Summary (Form IV VOA)
         GC/MS Instrument  Performance Check
           (Form V VOA)
         Internal Standard Area and RT Summary
           (Form VIII VOA)

     b.  Sample Data
         TCL Results - (Form I VOA)
         Tentatively Identified Compounds
           (Form I VOA-TIC)
         Reconstructed total ion chromatograms (RIC)
           for each sample
         For each sample:
           Raw spectra and background-subtracted
             mass spectra of target compounds
             identified
           Quantitation reports
           Mass spectra of all reported TICs with three
             best library matches

     c.  Standards  Data (All Instruments)
         Initial Calibration Data (Form VI VOA)
         RICs and Quan Reports for all Standards
         Continuing Calibration Data (Form VII VOA)
         RICs and Quantitation Reports for all Standards

    d.  Raw QC Data
         BFB
         Blank Data
         Matrix Spike/Matrix Spike Duplicate Data
                                          FORM DC-2-1

                                                                                  OLM01.7   7/91

-------
                    ORGANICS COMPLETE SDG FILE (CSF) INVENTORY SHEET (Cone.)
        CASE NO.
SDG NO.
SDG NOS.  TO FOLLOW
                               SAS NO.
                                                                    PAGE NOs
                                                                   FROM  TO
                                                       CHECK
                                                    LAB    EPA
 5.   Semivolatiles Data
     a.  QC Summary
          Surrogate Percent Recovery Summary (Form II SV)
          MS/MSD Summary (Form III SV)
          Method Blank Summary (Form IV SV)
          GC/MS Instrument Performance  Check
           (Form V SV)
          Internal Standard Area and RT Summary
           (Form VIII SV)

     b.  Sample Data
          TCL Results (Form I SV-1,  SV-2)
          Tentatively Identified Compounds (Form I SV-TIC)
          Reconstructed total ion chromatograms (RIC)
            for each sample
          For each sample:
            Raw spectra and background*subtracted
              mass spectra of target compounds
            QuantiCation reports
            Mass spectra of TICs with three  best library matches
            GPC chromatograms (if GPC performed)

     c.  Standards  Data (All Instruments)
          Initial Calibration Data (Form VI  SV-1,  SV-2)
          RICs and Quan Reports for  all Standards
          Continuing Calibration Data (Form  VII SV-1, SV-2)
          RICs and Quantitation Reports for  all Standards
          Semivolatile GPC Calibration  Data-UV
            detector traces

     d. Raw QC  Data
          DFTPP
          Blank Data
          Matrix Spike/Matrix Spike  Duplicate Data

6.   Pesticides

     a. QC Summary
          Surrogate  Percent Recovery Summary (Form II PEST)
          MS/MSD Duplicate  Summary (Form III PEST)
          Method Blank Summary (Form IV PEST)
                                          FORM DC-2-2
                                                                                  OLM01.7   7/91

-------
                  ORGANICS COMPLETE SDG FILE (CSF) INVENTORY SHEET ,(Cont.)
      CASE NO. 	  SDG NO.  	  SDG  NOS.  TO FOLLOW
                             SAS NO.                	
                                                                   PAGE NOs         CHECK
                                                                  FROM   TO      LAB    EPA
6.   Pesticides  (cont.)
     b.  Sample Data                                            	
         TCL Results - Organic Analysis Data Sheet
            (Form I PEST)
         Chromatograms (Primary Column)
         Chromatograins from second GC column confirmation
         GC Integration report or data system printout
         Manual work sheets
         For pesticides/Aroclors confirmed by GC/MS, copies
           of raw spectra and copies of background-subtracted mass
           spectra of target compounds (samples & standards)

     c.  Standards Data                                         	
         Initial Calibration of Single Component
           Analytes  (Form VI PEST-1 and PEST-2)
         Initial Calibration of Multicomponent Analytes
           (Form VI PEST-3)
         Analyte Resolution Summary (Form VI PEST-4)
         Calibration Verification Summary (Form VII PEST-1)
         Calibration Verification Summary (Form VII PEST-2)
         Analytical Sequence (Form VIII PEST)
         Florisil Cartridge Check (Form IX PEST-1)
         Pesticide GPC Calibration (Form IX PEST-2)
         Pesticide Identification Summary for Single Component
           Analytes (Form X PEST-1)
         Pesticide Identification Summary for Multicomponent
           Analytes (Form X PEST-2)
         Chromatograms and data system printouts
           A printout of retention times and corresponding peak
           areas or peak heights
         Pesticide GPC calibration data - UV detector traces

    d.  Raw QC  Data
         Blank Data                                            	
         Matrix Spike/Matrix Spike Duplicate Data
                                          FORM DC-2-3

                                                                                  OLM01.7   7/91

-------
                    ORGANICS COMPLETE SDG  FILE  (CSF) INVENTORY SHEET (Cont.)
        CASE NO.  	  SDG NO. 	  SDG NOS. TO FOLLOW
                              SAS NO.
                                                                    PAGE NOs        CHECK
                                                                   FROM  TO      LAB    EPA
 7.   Miscellaneous Data

      Original preparation  and analysis forms  or  copies  of
        preparation and analysis  logbook pages                 	  	   	   	
      Internal sample and sample  extract transfer
        chain-of-custody records                                	  	   	   	
      Screening records                                         	  	   	   	
      All  instrument output,  including strip charts
        from screening activities (describe or list)
 8.   EPA Shipping/Receiving Documents

      Airbills  (No. of shipments 	)
      Chain-of-Custody Records
      Sample Tags
      Sample Log-In Sheet  (Lab & DC1)
      Miscellaneous Shipping/Receiving Records
        (describe or list)
9.   Internal Lab Sample Transfer Records and Tracking Sheets
    (describe or list)
10.  Other Records (describe or list)

      Telephone Communication Log



11.  Comments:
Completed by:
  (CLP Lab)           (Signature)                   (Printed Name/Title)               (Date)
\udited by:     	  	  	
  (EPA)               (Signature)                  (Printed Name/Title)               (Date)
                                           FORM DC-2-4

                                                                                   OLM01.7   7/91

-------
                                  EXHIBIT C
                        TARGET COMPOUND LIST (TCL)  AND
                 CONTRACT REQUIRED QUANTITATION LIMITS (CRQL)
NOTE:  The values in these cables are quantitation limits,  not absolute
detection limits.  The amount of material necessary to produce a detector
response that can be identified and reliably quantified is  greater than that
needed to simply be detected above the background noise.  The quantitation
limits in these tables are set at the concentrations in the sample equivalent
to the concentratibn of the lowest calibration standard analyzed for each
analyte.

Specific quantitation limits are highly matrix dependent.  The quantitation
limits listed herein are provided for guidance and may not always be
achievable.

The CRQL values listed on the following pages are based on the analysis of
samples according the specifications given in Exhibit D.  For each fraction
and matrix, a brief synopsis of the sampling handling and analysis steps is
given, along with an example calculation for the CRQL value.  All CRQL values
are rounded to two significant figures.  For soil samples,  the moisture
content of the samples is not considered in these example calculations.
                              C-l       .                           OLM01.0

-------
  TARGET COMPOUND LIST (TCL) AND CONTRACT REQUIRED QUANTITATION LIMITS (CRQL)
Volatiles
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.
20.
21.
22.
23.
24.
25.
26.
27.
28.
29.
30.
31.
32.
33.
Chlorome thane
Bromome thane
Vinyl Chloride
Chloroe thane
Methylene Chloride
Acetone
Carbon Disulfide
1 , 1-Dichloroethene
1,1-Dichloroe thane
1,2-Dichloroethene (total)
Chloroform
1 , 2 -Dichloroe thane
2 - Butanone
1 , 1 , 1-Trichloroe thane
Carbon Tetrachloride
Bromodichlorome thane
1,2- Dichloropropane
cis-1, 3-Dichloropropene
Trichloroethene
Dibromochlorome thane
1 , 1 , 2-Trichloroethane
Benzene
trans -1,3- Dichloropropene
Bromoform
4 -Methyl - 2 -pentanone
2 - Hexanone
Tetrachloroethene
Toluene
1,1,2, 2 -Tetrachloroe thane
Chlorobenzene
Ethyl Benzene
S tyrene
Xylenes (Total)
CAS Number
74-87-3
74-83-9
75-01-4
75-00-3
75-09-2
67-64-1
75-15-0
75-35-4
75-34-3
540-59-0
67-66-3
107-06-2
78-93-3-
71-55-6
56-23-5
75-27-4
78-87-5
10061-01-5
79-01-6
124-48-1
79-00-5
71-43-2
10061-02-6
75-25-2
108-10-1
591-78-6
127-18-4
108-88-3
79-34-5
108-90-7
100-41-4
100-42-5
1330-20-7
Quant i
Water
ue/L
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
.tation
Low
Soil
ug/Kg
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
Limits*
Med.
Soil
ue/Ke
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
On
Column
(no.)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
* Quantitation limits listed for soil/sediment are based on wet weight.  The
quantitation limits calculated by the laboratory for soil/sediment,
calculated on dry weight basis as required by the contract, will be higher.
                                   C-2
OLM01.0

-------
Note  Chat the CRQL values listed on the preceding page may not be those
specified in previous CLP Statements of Work.  These values are set at
concentrations in the sample equivalent to the concentration of the lowest
calibration standard specified in Exhibit D VOA.  Lower quantitation limits
may be achievable for water samples by employing the Statement of Work for
Low Concentration Water for Organic Analyses.

VOLATILES

Water Samples

A 5 mL volume of water is purged with an inert gas at ambient temperature.
The volatiles are trapped on solid sorbents, and desorbed directly onto the
GC/MS.  For a sample with compound X at the CRQL of 10 ug/L:

(10 ug/L) (5 mL) (UT3  L/mL)  -  50 x  10"3 ug - 50 ng on the GC column

Low Level Soil/Sediment Samples

A 5 g aliquot of the soil/sediment sample is added to a volume of water in a
purge tube, heated, and purged with an inert gas.  The volatiles are trapped,
and later desorbed directly onto the GC/MS.  For a sample with compound X at
the CRQL of 10 ug/Kg:

(10 ug/Kg) (5 g) (10~3  Kg/g)  -  50 x  10*3 ug - 50 ng on the GC column

Medium Level Soil/Sediment Samples

A 4 g aliquot of soil/sediment is extracted with 10 mL of methanol, and
filtered through glass wool.  Only 1 mL of the methanol extract is taken for
screening and analysis.  Based on the results of a GC/FID screen, an aliquot
of the methanol extract is added to 5 mL of reagent water and purged at
ambient temperature.  The largest aliquot of extract considered in Exhibit D
is 100 uL.  For a sample with compound X at the CRQL of 1200 ug/Kg:

(1200 ug/Kg) (4 g) (ID'3 Kg/g)  - 4800 x  10'3 ug - 4800 ng

This  material is contained in the 10 mL methanol extract:

(4800 ng)/ 10 mL - 480 ng/mL

Of which,  100 uL are purged from the reagent water.

(480 ng/mL) (100 uL) (10~3 mL/uL) -  480  x  10"1 ng ) 50 ng on the GC column

Note  that for both low and medium soil/sediment samples, while it may affect
the purging efficiency, the volume of reagent water used in the purging
process does not affect the calculations.
                                    C-3                                OLM01.0

-------
 TARGET COMPOUND LIST  (TCL) AND CONTRACT REQUIRED QUANTITATION LIMITS (CRQL)
Semivolatiles
34.
35.
36.
37.
38.
39.
40.
41.

42.
43.

Phenol
bis(2-Chloroethyl) ether
2-Chlorophenol
1,3- Dichlorobenzene
1 , 4 -Dichlorobenzene
1,2- Dichlorobenzene
2-Methylphenol
2,2'-oxybis
( 1 - Chloropropane )
4-Methylphenol
N-Nitroso-di-n-
propylamine
CAS Number
108-95-2
111-44-4
95-57-8
541-73-T
106-46-7
95-50-1
95-48-7

108-60-1
106-44-5

621-64-7
Ouantitation Limits*
Low Med.
Water Soil Soil
ug/L ue/Ke ue/Ke
10
10
10
10
10
10
10

10
10

10
330
330
330
330
330
330
330

330
330

330
10000
10000
10000
10000
10000
10000
10000

10000
10000

10000
On
Column
Cne)
(20)
(20)
(20)
(20)
(20)
(20)
(20)

(20)
(20)

(20)
67-72-1
98-95-3
78-59-1
88-75-5
105-67-9
10
10
10
10
10
330
330
330
330
330
10000
10000
10000
10000
10000
(20)
(20)
(20)
(20)
(20)
44. Hexachloroethane
45. Nitrobenzene
46. Isophorone
47. 2-Nitrophenol
48. 2,4-Dimethylphenol

49. bis(2-Chloroethoxy)
     methane
50  2,4-Dichlorophenol
51. 1,2,4-Trichlorobenzene
52. Naphthalene
53. 4-Chloroaniline

54. Hexachlorobutadiene
55. 4-Chloro-3-methylphenol
56. 2-Methylnaphthalene
57. Hexachlorocyclopentadiene
58. 2,4,6-Trichlorophenol

59.. 2,4,5-Trichlorophenol
60. 2-Chloronaphthalene
61. 2-Nitroaniline
62. Dimethylphthalate
63. Acenaphthylene

64. 2,6-Dinitrotoluene
65. 3-Nitroaniline
66. Acenaphthene
67. 2,4-Dinitrophenol
68. 4-Nitrophenol
# Previously known by the name bis(2-Chloroisopropyl) ether
111-91-1
120-83-2
120-82-1
91-20-3
106-47-8
87-68-3
59-50-7
91-57-6
77-47-4
88-06-2
95-95-4
91-58-7
88-74-4
131-11-3
208-96-8
606-20-2
99-09-2
83-32-9
51-28-5
100-02-7
10
10
10
10
10
10
10
10
10
10
25
10
25
10
10
10
25
10
25
25
330
330
330
330
330
330
330
330
330
330
800
330
800
330
330
330
800
330
800
800
10000
10000
10000
10000
10000
10000
10000
10000
10000
10000
25000
10000
25000
10000
10000
10000
25000
10000
25000
25000
(20)
(20)
(20)
(20)
(20)
(20)
(20)
(20)
(20)
(20)
(50)
(20)
(50)
(20)
(20)
(20)
(50)
(20)
(50)
(50)
                                    C-4
OLM01.2  1/91

-------
    Semivolatiles
CAS Number
Ouantitation Limits*
       Low     Med.       On
Water  Soil    Soil    Column
ug/L   ug/Kg   ug/Kg	(ng)
800
330
330
330
330
25000
10000
10000
10000
10000
(50)
(20)
(20)
(20)
(20)
 69.  Dibenzofuran                 132-64-9     10      330    10000     (20)
 70.  2,4-Dinitrotoluene           121-14-2     10      330    10000     (20)
 71.  Diethylphthalate             84-66-2     10      330    10000     (20)
 72.  4-Chlorophenyl-phenyl
     ether                     7005-72-3     10      330    10000     (20)
 73.  Fluorene                     86-73-7     10      330    10000     (20)

 74.  4-Nitroaniline               100-01-6     25      800    25000     (50)
 75.  4,6-Dinitro-2-methylphenol   534-52-1     25      800    25000     (50)
 76.  N-nitrosodiphenylamine       86-30-6     10      330    10000     (20)
 77.  4-Bromophenyl-phenylether    101-55-3     10      330    10000     (20)
 78.  Hexachlorobenzene            118-74-1     10      330    10000     (20)

 79.  Pentachlorophenol            87-86-5     25
 80.  Phenanthrene                 85-01-8     10
 81.  Anthracene                   120-12-7     10
 82.  Carbazole                    86-74-8     10
 83.  Di-n-butylphthalate          84-74-2     10

 84.  Fluoranthene                 206-44-0     10      330    10000     (20)
 85.  Pyrene                       129-00-0     10      330    10000     (20)
 86.  Butylbenzylphthalate         85-68-7     10      330    10000     (20)
 87.  3,3'-Dichlorobenzidine       91-94-1     10      330    10000     (20)
 88.  Benzo(a)anthracene           56-55-3     10      330    10000     (20)

 89.  Chrysene                     218-01-9     10
 90. bis(2-Ethylhexyl)phthalate   117-81-7     10
 91.  Di-n-octylphthalate          117-84-0     10
 92.  Benzo(b)fluoranthene         205-99-2     10
 93.  Benzo(k)fluoranthene         207-08-9     10

 94.  Benzo(a)pyrene               50-32-8     10      330    10000     (20)
 95.  Indeno(l,2,3-cd)pyrene       193-39-5     10      330    10000     (20)
 96. Dibenz(a,h)anthracene        53-70-3     10      330    10000     (20)
 97.  Benzo(g,h,i)perylene         191-24-2     10      330    10000     (20)

* Quantitation limits listed for soil/sediment are based on wet weight.   The
quantitation limits calculated by the laboratory for soil/sediment,
calculated on dry weight basis as required by the contract, will be  higher.
330
330
330
330
330
10000
10000
10000
10000
10000
(20)
(20)
(20)
(20)
(20)
                                    C-5
                                OLM01.1  12/90

-------
SEMIVOLATILES

Water Samples

AIL volume of water is extracted in a continuous liquid-liquid extractor
with methylene chloride at a pH of approximately 2.  This extract is reduced
in volume to 1.0 mL, and a 2 uL volume is injected onto the GC/MS for
analysis.  For a sample with compound X at the CRQL of 10 ug/L:

(10 ug/L) (1 L) - 10 ug in the original extract

When the extract is concentrated, this material is contained in the 1 mL
concentrated extract, of which 2 uL are injected into the instrument:

(10 ug/mL) (2 uL) (10"3  mL/uL)  -  20  x  10'3 ug - 20 ng on the GC column

Low Soil Samples

A 30 g soil sample is extracted three times with methylene chloride/acetone
at ambient pH, by sonication.  The extract is reduced in volume to 1.0 mL,
and a 2 uL volume is injected onto the GC/MS for analysis.  For a sample with
compound X at the CRQL of 330 ug/Kg:

(330 ug/Kg) (30 g) (10-3 Kg/g)  -  9900  x  10'3 ug - 9.9 ug

When the sample extract is to be subjected to Gel Permeation Chromatography
(required) to remove high molecular weight interferences, the volume of the
extract is initially reduced to 10 mL.  This 10 mL is put through the GPC
column, and only 5 mL are collected off the GPC.  That 5 mL volume is reduced
to 0.5 mL prior to analysis.  Therefore:

(9.9 ug/10 mL) (5 mL) - 4.95 ug

This material is contained in the 0.5 mL extract, of which 2 uL are injected
into the instrument:

(4.95 ug/0.5 mL) (2 uL) (10'3 mL/uL) - 1.98 x  10'2 ug ) 20 ng on the GC column

Medium Soil Samples

A 1 g soil sample is extracted once with 10 mL of methylene chloride/acetone,
which is filtered through glass wool to remove particles of soil.  The
filtered extract is then subjected to GPC clean up, and only 5 mL of extract
are collected after GPC.  This extract is reduced in volume to 0.5 mL, of
which 2 uL are injected onto the GC/MS.   For a sample with compound X at the
CRQL of 10,000 ug/Kg:
                                     l
(10,000 ug/Kg) (Ig) (lO'3 Kg/g) - 10 ug
(continued)
                                    C-6                          OLM01.2  1/91

-------
Semivolatiles, Medium Soil, continued -

This material is contained in the 10 mL extract, of which only 5 mL are
collected after GPC:

(10 ug) (5 mL/10mL) - 5 ug

The volume of this extract is reduced to 0.5 mL, of which 2 uL are injected
into the instrument:

(5 ug/0.5 mL) (2 uL) (10'3  mL/uL) -  20 x 10'3 ug - 20 ng on the GC column

Eight semivolatile compounds are calibrated using only a four point initial
calibration, with the lowest standard at 50 ng.  Therefore, the CRQL values
for these eight compounds are 2.5 times higher for all matrices and levels.
                                   C-7                         OLM01.1  12/90

-------
  TARGET COMPOUND LIST (TCL)  AND CONTRACT REQUIRED QUANTITATION LIMITS  (CRQL)
   Pesticides/Aroclors
 CAS Number
Ouantitation Limits*
Water   Soil    On Column
ug/L    ug/Ke	(pg)
98.
99.
100.
101.
102.
alpha -BHC
beta-BHC
delta -BHC
gamma -BHC (Lindane)
Heptachlor
319-84-6
319-85-7
319-86-8
58-89-9
76-44-8
0.05
0.05
0.05
0.05
0.05
1.7
1.7
1.7
1.7
1.7
5
5
5
5
5
103. Aldrin
104. Heptachlor epoxide
105. Endosulfan I
106. Dieldrin
107. 4,4'-DDE

108. Endrin
109. Endosulfan II
110. 4,4'-ODD
111. Endosulfan sulfate
112. 4,4'-DDT

113. Me thoxychlor
114. Endrin ketone
115. Endrin aldehyde
116. alpha-Chlordane
117. gamma-Chlordane

118. Toxaphene
119. Aroclor-1016
120. Aroclor-1221
121. Aroclor-1232
122. Aroclor-1242

123. Aroclor-1248
124. Aroclor-1254
125. Aroclor-1260
309-00-2
1024-57-3
959-98-8
60-57-1
72-55-9
0.05
0.05
0.05
0.10
0.10
1.7
1.7
1.7
3.3
3.3
72-20-8
33213-65-9
72-54-8
1031-07-8
50-29-3
0.10
0.10
0.10
0.10
0.10
3.3
3.3
3.3
3.3
3.3
   72-43-5
53494-70-5
7421-36-3
 5103-71-9
 5103-74-2

 8001-35-2
12674-11-2
11104-28-2
11141-16-5
53469-21-9

12672-29-6
11097-69-1
11096-82-5
0.50
0.10
0.10
0.05
0.05
2.0
1.0
1.0
1.0
1.0
1.0
 17.0
  3.3
  3.3
  1.7
  1.7

170.0
 33.0
 67.0
 33.0
 33.0

 33.0
 33.0
 33.0
  5
  5
  5
 10
 10

 10
 10
 10
 10
 10

 50
 10
 10
  5
  5

500
100
200
100
100

100
100
100
* Quantitation limits listed for soil/sediment are based on wet weight.  The
quantitation limits calculated by the laboratory for soil/sediment,
calculated on dry weight basis as required by the contract, will be higher.

There is no differentiation between the preparation of low and medium  soil
samples in this method for the analysis of Pesticides/Aroclors.
                                    C-8
                                 OLM01.1  12/90

-------
 PESTICIDES/AROCLORS

 Water  Samples

 AIL  volume of water  is extracted three times with methylene chloride or by
 a  continuous liquid-liquid extractor.  This extract is reduced in volume to
 approximately  3-5 mL,  and diluted up to 10.0 mL with clean solvent.  When Gel
 Permeation Chromatography is performed, only 5 of the 10 mL of extract are
 collected after GPC.

 Regardless of  whether  GPC is performed, either 1.0 or 2.0 mL of the 10.0 mL
 of the original extracts are taken through the remaining clean up steps
 (Florisil and  sulfur removal).  The volume taken through Florisil cleanup and
 the  final volume of the extract after the clean up steps depends on the
 requirements of the autosampler.  If the autosampler can handle 1.0 mL final
 extract volumes, this  is the volume taken through Florisil and the final
 volume.  If the autosampler cannot reliably handle 1.0 mL volumes, the volume
 is 2.0 mL.  When using an autosarapler, the injection volume may be 1.0 or 2.0
 uL.  Manual injections must use a 2.0 uL injection volume.

 For  a  sample with compound X at the CRQL of 0.05 ug/L and an autosampler
 requiring a 1.0 mL volume:

 (0.05  ug/L) (1 L) - 0.05 ug in the original extract

 This material  is contained in the 10.0 mL of extract:

 (0.05  ug)/ (10.0 mL) - 0.005 ug/raL
                                              •
 Of which, only 1.0 mL  is carried through the remaining clean up steps.  For a
 final  extract  volume of 1.0 mL and a 1 uL injection volume:

 (0.005 ug/raL) (1 uL)(KT3 mL/uL)  -  5 x  10'6 ug - 5 pg on the GC column

 Soil Samples

 There  is no differentiation between the preparation of low and medium soil
 samples in this method for the analysis of pesticides/Aroclors.  A 30 g soil
 sample is extracted three times with methylene chloride/acetone by
 sonication.  The extract is reduced in volume to 10.0 mL and subjected to Gel
 Permeation Chromatography.  After GPC, only 5.0 mL of extract are collected.
 However, as with the water sample described above, either 1.0  or 2.0 mL of
 that extract are subjected to the other clean up steps, so no loss of
 sensitivity results from the use of GPC.  From this point on, the soil sample
 extract is handled in  the same fashion as the extract of a water sample.  For
 a sample with  compound X at the CRQL of 1.7 ug/Kg:

 (1.7 ug/Kg) (30 g) (10'3 Kg/g) - 51 x  10'3 ug - 51 ng in the original extract

This material  is contained in the 10.0 mL of extract:

 (51 ng)/ 10 mL - 5.1 ng/mL

 (continued)


                                   C-9                          OLM01.2  1/91

-------
Pesticides/Aroclors, continued -

Of which, only 1.0 or 2.0 mL are carried through the remaining cleanup steps.
For a final extract volume of 1.0 mL and a 1 uL injection volume:

(5.1 ng/n>L)(l uLKlO'3 mL/uL) -  5.1 x 10'3 ng ) 5 pg on the GC column.

For either water or soil samples, if the autosampler used requires a 2.0 mL
final volume, the concentration in the 10.0 mL of extract above remains the
same.

Using a 2 uL injection volume,  twice the total number of picograms are
injected onto the GC column.  However,  because the injection volume must be
the same for samples and standards, twice as much material is injected onto
the column during calibration,  and thus the amount of compound X injected
from the sample extract is equivalent to the amount of compound X injected
from the calibration standard,  regardless of injection volume.

If a single injection is used for two GC columns attached to a single
injection port, it may be necessary to use an injection volume greater than 2
uL.
                                   C-10                         OLM01.5  4/91

-------
     EXHIBIT  D
ANALYTICAL METHODS
   FOR VOLATILES
   D-l/VOA                             OLM01.0

-------
                              Table of Contents


                                                                 Page

SECTION I -    Introduction 	  D-3/VOA

SECTION II -   Sample Preparation and Storage 	  D-5/VOA

               PART A -  Sample Storage and Holding Times 	  D-6/VOA

               PART B -  Protocols for Hexadecane
                        Extraction of Volatiles from
                        Water and Soil/Sediment for
                        Optional Screening 	  D-7/VOA

SECTION III -  Optional Screening of Hexadecane
               Extracts for Volatiles 	  D-10/VOA

SECTION IV -   GC/MS Analysis of Volatiles 	  D-14/VOA
                                 D-2/VOA                              OLM01.0

-------
                                  SECTION I
                                 INTRODUCTION
The analytical methods that follow are designed to analyze water, sediment
and soil from hazardous waste sites for the organic compounds on the Target
Compounds List (TCL, see Exhibit C).   The methods are based on EPA Method 624
(Purgeables).

The methods are divided into the following sections:  sample preparation,
screening,  and analysis.  Sample preparation covers sample storage, sample
holding times, and medium level sample extraction.  As described in the
screening section, a portion of a hexadecane extract may be screened on a gas
chromatograph with appropriate detectors to determine the concentration level
of organics.  The analysis section contains the GC/MS analytical methods for
organics.  The purge and trap technique, including related sample
preparation, is included in the analysis section because GC/MS operation and
the purge and trap technique are interrelated.
                                 D-3/VOA                              OLM01.0

-------
                                                               SECTION I


1.     Method for the Determination of Volatile (Purgeable) Organic Compounds

1.1   Scope and Application

      This method covers the determination of the target volatile (purgeable)
      organics as listed in Exhibit C.   The contract required quantitation
      limits are also listed in Exhibit C.   The method includes an optional
      hexadecane screening procedure.  The extract is screened on a gas
      chromatograph/flame ionization detector (GC/FID) to determine the
      approximate concentration of organic constituents in the sample.  The
      actual analysis is based on a purge and trap gas chromatographic/mass
      spectrometer (GC/MS) method.  For soil/sediment samples, the purge
      device is heated.

1.2   Problems have been associated with the following compounds analyzed by
      this method:

          o   Chloromethane, Vinyl chloride, Bromomethane, and Chloroethane
             can display peak broadening if the compounds are not delivered
             to the GC column in a tight band.

          o   Acetone, Hexanone,  2-Butanone, and 4-methyl-2-Pentanone have
             poor purge efficiencies.

          o   1,1,1-Trichloroethane and all the Dichloroethanes can
             dehydrogenate during storage or analysis.

          o   Tetrachloroethane and 1,1-Dichloroethane can be degraded by
             contaminated transfer lines in purge and trap systems and/or
             active sites in trapping materials.

          o   Chloromethane can'be lost if the purge flow is too fast.
          o   Bromoform is one of the compounds most likely to be adversely
             affected by cold spots and/or active sites in the transfer
             lines.  Response of its quantitation ion (m/z 173) is directly
             affected by the tuning of the GC/MS to meet the instrument
             performance criteria for BFB at ions m/z 174/176.  Increasing
             the m/z 174/176 ratio may improve Bromoform response.
                                 D-4/VOA                             OLM01.0

-------
          SECTION II
SAMPLE PREPARATION AND STORAGE
         D-5/VOA                             OLM01.0

-------
                                                              SECTION II
PART A - SAMPLE STORAGE AND HOLDING TIMES

1.     Procedures for Sample Storage

      The samples must be protected from light and refrigerated at 4°C (±2°C)
      from the time of receipt until 60 days after delivery of a complete
      sample data package to the Agency.   After 60 days,  the samples may be
      disposed of in a manner that complies with all applicable regulations.

      The samples must be stored in an atmosphere demonstrated to be free of
      all potential contaminants and in a refrigerator used only for storage
      of purgeable samples received under this contract.

      Samples, sample extracts,  and standards must be stored separately.

      Volatiles standards must be stored separately from semivolatile and
      pesticide/Aroclor standards.

2.     Contract Required Holding Times

      Analysis of water and soil/sediment samples must be completed within 10
      days of validated time of sample receipt (VTSR).
                                 D-6/VOA                        OLM01.2   1/91

-------
PART B -  PROTOCOLS  FOR  HEXADECANE  EXTRACTION OF VOLATILES FROM WATER AND
          SOIL/SEDIMENT  FOR OPTIONAL  SCREENING

1.     Summary of Method

1.1   Matrices

      1.1.1   Water - a 40  mL aliquot of sample is extracted with 2 mL of
              hexadecane.   This provides a minimum quantitation limit (MQL)
              as  follows:

              Compounds                             MQL ug/L
              non-halogenated aromatics              40-  50
              halogenated methanes                  800-1000
              halogenated ethanes                   400- 500

      1.1.2   Soil/sediment - 40 mL  of reagent water are added to 10 g (wet
              weight) of soil and shaken.  The water phase is in turn
              extracted with 2 mL of hexadecane.  This provides a minimum
              quantitation  limit of  approximately four times higher than
              those listed  for water.

1.2   The hexadecane extraction and screening protocols for  purgeables  are
      optional.  These protocols are included to aid the analyst  in deciding
      whether a sample is low or medium level.   The use of these or other
      screening protocols could prevent saturation of the purge  and trap
      system and/or the GC/MS system.  It is recommended that these or  other
      screening protocols be used,  particularly if there is  some doubt  about
      the level of organics in a sample.   This is  especially true in
      soil/sediment analysis.

2.     Limitations

      These extraction and preparation procedures  were developed for rapid
      screening of water samples from hazardous waste sites.   The design  of
      the methods thus does not stress efficient recoveries  or low limits of
      quantitation.   Rather, the procedures were designed to screen at
      moderate recovery and sufficient sensitivity for a broad spectrum of
      organic chemicals.  The results of the analyses thus may reflect  only a
      minimum of the amount actually present in some samples.   This is
      especially true if water soluble solvents are present.

3.     Interferences

3.1   Method interferences may be caused by contaminants in solvents,
      reagents,  glassware,  and other sample processing hardware  that lead to
      discrete artifacts and/or elevated baselines in the total  ion current
      profiles.   All of these materials must be routinely demonstrated  to be
      free from interferences under the conditions of the analysis by running
      laboratory reagent blanks.   Matrix interferences may be caused by
      contaminants that are coextracted from the sample.  The extent of
      matrix interferences will vary considerably  from source to source,
      depending upon the nature and diversity of the site being  sampled.


                                 D-7/VOA                             OLM01.0

-------
4.     Apparatus and Materials

4.1   Vials and caps - 2 mL for GC -auto sampler.

4.2   Volumetric flask - 50 mL with  ground glass  stopper.

4.3   Pasteur pipets - disposable.

4.4   Centrifuge tube - 50 mL with ground glass stopper or Teflon-lined screw
      cap.

4.5   Balance - analytical, capable  of accurately weighing + 0.0001 g.

5.     Reagents

5.1   Hexadecane and methanol - pesticide residue analysis grade or
      equivalent.

5.2   Reagent water - defined as water in which an interferent is not
      observed at the CRQL of each parameter of interest.

5.3   Standard mixture #1 containing benzene,  toluene,  ethyl benzene and
      xylene.  Standard mixture #2 containing n-nonane  and n-dodecane.

      5.3.1    Stock standard solutions (1.00 ug/uL) - prepared  from pure
               standard materials or purchased as  certified solutions.

               5.3.1.1    Prepare stock standard solutions by accurately
                         weighing about 0.0100 g  of pure material.   Dissolve
                         the material in methanol and dilute to volume in a
                         10 mL volumetric flask.   Larger volumes can be used
                         at the convenience of the analyst.  If compound
                         purity is certified at 97% or greater, the weight
                         can be used  without correction to calculate the
                         concentration of the stock standard.

               5.3.1.2    Transfer the stock standard solutions into multiple
                         Teflon-sealed screw-cap  vials.  Store, with no
                         head-space, at -10°C to  -20°C, and protect from
                         light.  Stock standard solutions should be checked
                         frequently for signs of  degradation or evaporation,
                         especially just prior to preparing calibration
                         standards from them.  These solutions must be
                         replaced after six months, or sooner,  if comparison
                         with quality control check samples indicates a
                         problem.  Standards prepared from gases or reactive
                         compounds such as styrene must be replaced after two
                         months, or sooner, if comparison with  quality
                         control check samples indicates a problem.

      5.3.2    Prepare working  standards of mixtures #1 and #2  at 100 ng/uL  of
               each compound  in methanol.  Store  these solutions as  in  5.3.1.2
               above.
                                 D-8/VOA                        OLM01.2   1/91

-------
6.    Sample Extraction

6.1   Water

      6.1.1   Allow the contents of the 40 mL sample vial to come to room
              temperature.  Quickly transfer the contents of the 40 mL sample
              vial to a 50 mL volumetric flask.  Immediately add 2.0 mL of
              hexadecane, cap the flask, and shake vigorously for 1 minute.
              Let phases separate.  Open the flask and add sufficient reagent
              water to bring the hexadecane layer into the neck of the flask.

      6.1.2   Transfer approximately 1 mL of the hexadecane layer to a 2.0 mL
              GC vial.  If an emulsion is present after shaking the sample,
              break it by doing the following:

              o    Pulling the emulsion through a small plug of Pyrex glass
                   wool packed in a pipet,  or

              o    Transferring the emulsion to a centrifuge tube and
                   centrifuging for several minutes.

      6.1.3   Add 200 uL each of working standard #1 and #2 to separate 40 mL
              portions of reagent water in 50 mL volumetric flasks.  Follow
              steps 6.1.1 - 6.1.2, starting with the immediate addition of
              2.0 mL of hexadecane.

6.2   Soil/Sediment

      6.2.1   Add approximately 10 g of soil (wet weight) to 40 mL of reagent'
              water in a 50 mL centrifuge tube with a ground glass stopper or
              teflon-lined cap.  Cap and shake vigorously for one minute.
              Centrifuge the capped flask briefly.  Quickly transfer
              supernatant water to a 50 mL volumetric flask equipped with a
              ground-glass stopper.

      6.2.2   Follow 6.1, starting with the addition of 2.0 mL of hexadecane.

7.    Sample Analysis

      The sample is ready for GC/FID screening.   Proceed to Section III,
      Optional Screening of Hexadecane Extracts for Volatiles.
                                 D-9/VOA                        OLM01.2   1/91

-------
           SECTION  III
OPTIONAL SCREENING OF HEXADECANE
     EXTRACTS FOR VOLATILES
          D-10/VOA                             OLM01.0

-------
                                                             SECTION III
 1.    Summary of Method

      The hexadecane extracts of water and soil/sediment are screened on a
      gas chromatograph/flame ionization detector (GC/FID).   The results of
      the screen will determine if volatile organics are to be analyzed by
      low or medium level GC/MS procedures if the sample is a soil/sediment,
      or to determine the appropriate dilution factor if the sample is water.

 2.    Apparatus and Materials

 2.1   Gas chromatograph - an analytical system complete with a temperature
      programmable gas chromatograph suitable for on-column injection and all
      required accessories including syringes, analytical columns, gases,
      detector, and strip-chart recorder.  A data system is recommended for
      measuring peak areas.

      2.1.1   Above-described GC, equipped with  flame  ionization  detector.

      2.1.2   GC  column  - 3 m x 2 mm ID  glass column packed with  10% OV-101
              on  100-120 mesh Chromosorb W-HP (or equivalent).  The column
              temperature should be programmed from 80"C  to 280°C  at
              16°C/min. and held at 280°C for 10 minutes.

 3.    Reagents

      Hexadecane - pesticide residue analysis grade or equivalent.

4.    Limitations

4.1   The flame ionization detector varies considerably in sensitivity when
      comparing aromatics and halogenated methanes and ethanes.  Halomethanes
      are approximately 20x less sensitive than aromatics and haloethanes are
      approximately lOx less sensitive than aromatics.   Low molecular weight,
      water soluble solvents,  e.g.,  alcohols and ketones, will not extract
      from the water,  and therefore will not be detected by the GC/FID.

4.2   Following are two options for interpreting the GC/FID chromatogram.

      4.2.1   Option A is to use standard mixture #1 containing the aromatics
              to calculate an approximate concentration  of the aromatics  in
              the  sample.  Use this information  to determine  the  proper
              dilution for purge and trap if the sample  is a  water, or
              whether to use the low or medium level GC/MS purge  and trap
              methods if the sample is a soil/sediment  (see Table  1,
              paragraph 6.2.1.3 for guidance).  This should be the best
              approach; however, the aromatics may be  absent  or obscured by
              higher concentrations of other purgeables.  In  these cases,
              Option B may be the best approach.
                                 D-ll/VOA                       OLM01.2   1/91

-------
      4.2.2   Option B  is to use standard mixture #2 containing n-nonane and
              n-dodecane to calculate a factor.  Use the factor to calculate
              a dilution for purge and trap of a water sample or to determine
              whether to use the low or medium level GC/MS purge and trap
              methods for soil/sediment samples (see Table 1, paragraph
              6.2.1.3 for guidance).  All purgeables of interest have
              retention times less than the n-dodecane.

5.     Extract Screening

5.1   External standard calibration -  Standardize  the GC/FID each 12 hr.
      shift for half scale  response.   This is done by injecting 1-5 uL of the
      extracts that contain approximately 10 ng/uL each of mix #1 and mix #2
      compounds,  as prepared in 5.3,  Section II,  Part B.   Use the GC
      conditions specified  in paragraph 2.1.2.

5.2   Inject the same volume of hexadecane extract as the  extracted standard
      mixture in 5.1.    Use the GC conditions specified in 2.1.2.

6.     Analytical Decision Point

6.1   Water

      6.1.1   Compare the chromatograms of the hexadecane extract of the
              sample with those of the reagent blank and extract of the
              standard.

              6.1.1.1    If no peaks are noted, other than those also in the
                         reagent blank, analyze a 5 mL water sample by purge
                         and trap GC/MS.

              6.1.1.2    If peaks are present prior to the n-dodecane and the
                         aromatics are distinguishable, follow Option A
                         (4.2.1).

              6.1.1.3    If peaks are present prior to the n-dodecane but the
                         aromatics are absent or indistinguishable, use
                         Option B as follows: if all peaks are <3% of the
                         n-nonane, analyze a 5 mL water sample by purge and
                         trap GC/MS.  If any peaks are >3% of the n-nonane,
                         measure the peak height or area of the major peak
                         and calculate the dilution factor as follows:

                         peak area of sample major peak x 50 = dilution
                             peak area of n-nonane             factor

                         The water sample will be diluted using the
                         calculated factor just prior to purge and trap GC/MS
                         analysis.
                                 D-12/VOA                             OLM01.0

-------
6.2   Soil/Sediment

      6.2.1   Compare  the chromatograms of the hexadecane extract of the
              sample with those of the reagent blank and extract of the
              standard.

              6.2.1.1    If no peaks are noted, other than those also in the
                         reagent blank, analyze a 5 g sample by low level
                         GC/MS.

              6.2.1.2    If peaks are present prior to the n-dodecane and the
                         aromatics are distinguishable, follow Option A
                         (4.2.1) and the concentration information in Table
                         1, paragraph 6.2.1.3, to determine whether to
                         analyze by low or medium level method.

              6.2.1.3    If peaks are present prior to the n-dodecane but the
                         aromatics are absent or indistinguishable, and using
                         Option B as follows, calculate a factor using the
                         following formula:

                         peak area of sample major peak =• X Factor
                              peak area of n-nonane
      Table 1  -  Determination of GC/MS Purge & Trap Method

                                                         Approximate
                                        '            Concentration Range*
        X Factor          Analyze by     ;         	Cue/kg)	

          0-1.0         low level method                   0-1,000
          >1.0         medium level method                >1,000

    This  concentration range  is  based  on the  response  of aromatics  to GC/FID.
    When  comparing GC/FID responses, the concentration for halomethanes is 20x
    higher,  and  that for haloethanes is  lOx higher.
6.3   Sample Analysis

      Proceed to Section IV, GC/MS Analysis of Volatiles.
                                 D-13/VOA                             OLM01.0

-------
  SECTION IV
GC/MS ANALYSIS
 OF VOLATILES
                                     OLMOI.O

-------
                                                              SECTION IV
1.    Summary of Methods

1.1   Water Samples

      An inert gas is bubbled through a 5 mL sample contained in a
      specifically designed purging chamber at ambient temperature.   The
      purgeables are efficiently transferred from the aqueous phase to the
      vapor phase.  The vapor is swept through a sorbent column where the
      purgeables are trapped. After purging is completed,  the sorbent column
      is heated and backflushed with the inert gas to desorb the purgeables
      onto a gas chromatographic column.  The gas chromatograph is
      temperature programmed to separate the purgeables which are then
      detected with a mass spectrometer.

      An aliquot of the sample is diluted with reagent water when dilution is
      necessary.  A 5 mL aliquot of the dilution is taken for purging.

1.2   Soil/Sediment Samples

      1.2.1   Low  level  -  an inert gas  is bubbled  through a mixture of
              reagent water  and  5 g of  sample contained in a specifically
              designed purging chamber  that is held at an elevated
              temperature.   The  purgeables are efficiently transferred  from
              the  aqueous  phase  to the vapor phase.  The vapor  is swept
              through a  sorbent  column where the purgeables are  trapped.
              After purging  is completed, the sorbent column is  heated  and
              backflushed  with the inert gas to desorb the purgeables onto a
              gas  chromatographic column.  The gas chromatograph is
              temperature  programmed to separate the purgeables, which  are
              then detected  with a mass spectrometer.

      1.2.2   Medium level -  a measured amount of  soil is extracted with
              methanol.  A portion of the methanol extract is diluted to 5 mL
              with reagent water.  An inert gas is bubbled through this
              solution in  a  specifically designed  purging chamber at ambient
              temperature.   The  purgeables are effectively transferred  from
              the  aqueous  phase  to the vapor phase.  The vapor  is swept
              through a  sorbent  column where the purgeables are  trapped.
              After purging  is completed, the sorbent column is  heated  and
              backflushed  with the inert gas to desorb the purgeables onto a
              gas  chromatographic column.  The gas chromatograph is
              temperature  programmed to separate the purgeables, which  are
              then detected  with a mass spectrometer.

2.    Interferences

2.1   Impurities in the purge gas,  organic compounds out-gassing from the
      plumbing ahead of the trap, and solvent vapors in the laboratory
      account for the majority of contamination problems.   The analytical
      system must be demonstrated to be free from contamination under the
      conditions of the analysis by running laboratory reagent blanks as
      described in Exhibit E.  The use of non-TFE tubing,  non-TFE thread
      sealants,  or flow controllers with rubber components in the purging
      device should be avoided.

                                 D-15/VOA                       OLM01.2   1/91

-------
                                                              SECTION IV
2.2   Samples can be contaminated by diffusion of volatile organics
      (particularly fluorocarbons and methylene chloride) through the septum
      seal into the sample during storage and handling.  A holding blank
      prepared from reagent water and carried through the holding period and
      the analysis protocol serves as a check on such contamination.  One
      holding blank per case must be analyzed.  Data must be retained by the
      laboratory and be made available for inspection during on-site
      laboratory evaluations.

2.3   Contamination by carry-over can occur whenever high level and low level
      samples are sequentially analyzed.  To reduce carryover, the purging
      device and sampling syringe must be rinsed with reagent water between
      sample analyses.   Whenever an unusually concentrated sample is
      encountered, it must be followed by an analysis of reagent water to
      check for cross contamination.  For samples containing large amounts of
      water soluble materials, suspended solids, high boiling compounds or
      high purgeable levels, it may be necessary to wash out the purging
      device with a detergent solution between analyses, rinse it with
      distilled water,  and then dry it in a 105°C oven.  The trap and other
      parts of the system are also subject to contamination; therefore,
      frequent bakeout and purging of the entire system may be required.

2.4   The laboratory where volatile analysis is performed should be
      completely free of solvents.

3.     Apparatus and Materials

 j    Brand names, suppliers,  and part numbers are for illustrative purposes
      only.  No endorsement is implied.  Equivalent performance may be
   /  achieved using apparatus and materials other than those specified here,
      but demonstration of equivalent performance meeting the requirements of
      this SOW is the responsibility of the Contractor.

3.1   Micro syringes -  25 uL and larger, 0.006 inch ID needle.

3.2   Syringe valve - two-way, with Luer ends (three each), if applicable to
      the purging device.

3.3   Syringe - 5 mL, gas-tight with shut-off valve.

3.4   Balance - analytical, capable of accurately weighing + 0.0001 g, and a
      top-loading balance capable of weighing + 0.1 g

3.5   Glassware

      3.5.1 Bottle - 15 mL, screw cap,  with Teflon cap liner.

      3.5.2 Volumetric  flasks  - class A with ground-glass stoppers.

      3.5.3 Vials - 2 mL for GC autosampler.

3.6   Purge and trap device -  consists of three separate pieces of equipment:
      the sample purger,  trap, and the desorber.  Several complete devices
      are now commercially available.

                                 D-16/VOA                       OLM01.2  1/91

-------
                                                              SECTION IV
      3.6.1   The sample purger must be designed to accept 5 mL samples with
              a water column at least 3 cm deep.  The gaseous head space
              between the water column and the trap must have a total volume
              of less than 15 mL.  The purge gas must pass through the water
              column as finely divided bubbles, each with a diameter of less
              than 3 mm at the origin.  The purge gas must be introduced no
              more than 5 mm from the base of the water column.

      3.6.2   The trap must be at least 25 cm long and have an inside
              diameter of at least 0.105 inch.  The trap must be packed to
              contain the following minimum lengths of absorbents: 15 cm of
              2,6-diphenylene oxide polymer (Tenax-GC, 60/80 mesh) and 8 cm
              of silica gel (Davison Chemical, 35/60 mesh, grade 15, or
              equivalent).

      3.6.3   The desorber should be capable of rapidly heating the trap to
              180°C.  The polymer section of the trap should not be heated
              higher than 180°C and the remaining sections should not exceed
              220°C during bakeout mode.

      3.6.4   The purge and trap device may be assembled as a separate unit
              or be coupled to a gas chromatograph.

      3.6.5   A heater or heated bath capable of maintaining the purge device
              at 40"C + 1°C is to be used for low level soil analysis, but
              not for waters or medium level soil analyses.

3.7   GC/MS system

      3.7.1   Gas Chromatograph - the gas chromatograph (GC) system must be
              capable of temperature programming and have a flow controller
              that maintains a constant column flow rate throughout
              desorption and temperature program operations.  The system must
              include or be interfaced to a purge and trap system as
              specified in paragraph 3.6 and have all required accessories
              including syringes, analytical columns, and gases.  All GC
              carrier gas lines must be constructed from stainless steel or
              copper tubing.  Non-polytetrafluoroethylene (PTFE) thread
              sealants, or flow controllers with rubber components are not to
              be used. If capillary columns are to be used (see below), the
              column oven must be cooled to 10°C; therefore, a subambient
              oven controller is required.

      3.7.2   Gas Chromatography Columns

              3.7.2.1    Packed Columns - 6 ft long x 0.1 in ID glass, packed
                         with 1% SP-1000 on Carbopack B (60/80 mesh) or
                         equivalent.
                                 D-17/VOA                       OLM01.2   1/91

-------
                                                        SECTION IV
                   NOTE:  Capillary columns may be used for analysis of
                   volatiles, as long as the Contractor uses the
                   instrumental parameters in EPA Method 524.2 as
                   guidelines, uses the internal standards and
                   surrogates specified in this contract, and
                   demonstrates that the analysis meets all of the
                   performance and QA/QC criteria contained in this
                   contract.

         3.7.2.2    Capillary Columns

                   o  30 m long x 0.53 mm ID VOCOL  (Supelco, Inc.,  or
                      equivalent) fused  silica wide-bore  capillary
                      column with 3 urn film thickness.

                       OR

                   o  30 m long x 0.53 mm ID DB-624 fused silica  wide-
                      bore (J&W Scientific, Inc., or equivalent)  column
                      with 3 urn film thickness.

3.7.3    Mass Spectrometer  -  must be capable of scanning  from 35  to  300
         amu every 1 second  or less, utilizing 70 volts (nominal)
         electron energy  in  the electron impact ionization mode,  and
         producing a mass  spectrum which meets all  the instrument
         performance acceptance criteria when 50 ng of
         p-bromofluorobenzene  (BFB) is injected through the gas
         chromatograph inlet.  The instrument conditions  required for
         the acquisition  of  the BFB mass spectrum are given in paragraph
         6.4.3.

         NOTE:  BFB criteria must be met before any sample extracts  are
         analyzed.  Any samples analyzed when BFB criteria have not  been
         met will require  reanalysis at no cost to  the Agency.  To
         ensure sufficient precision of mass spectral data, the MS  scan
         rate should allow acquisition of at least  five spectra while a
         sample compound  elutes from the GC.  The purge and trap  GC/MS
         system must be in a room whose atmosphere  is demonstrated  to be
         free of all potential contaminants which will interfere with
         the analysis.  The  instrument must be vented to  the outside of
         the facility  or  to  a trapping system which prevents the  release
         of contaminants  into the instrument room.

3.7.4    GC/MS interface  - any gas chromatograph to mass  spectrometer
         interface that gives acceptable calibration points, at 50  ng or
         less per injection,  for each of the parameters of interest  and
         achieves all  acceptance criteria may be used.  Gas
         chromatograph to  mass spectrometer interfaces constructed  of
         all-glass or  glass-lined materials are recommended. Glass  can
         be deactivated by silanizing with dichlorodimethylsilane.
                           D-18/VOA                       OLM01.2   1/91

-------
                                                              SECTION IV


      3.7.5    Data  system  -  a  computer system must be  interfaced to  the mass
               spectrometer that allows the continuous  acquisition and
               storage, on machine readable media, of all tnass spectra
               obtained throughout the duration of the  chromatographic
               program.  The  computer must have software that allows
               searching any  GC/MS data file for ions of a specified  mass and
               plotting such  ion abundances versus time or scan number.  This
               type  of plot is  defined as an Extracted  Ion Current Profile
               (EICP).  Software must also be available that allows
               integrating the  abundance in any EICP between specified  time
               or  scan number limits.  Also, for the non-target compounds,
               software must  be available that allows for the comparison of
               sample spectra against reference library spectra.  The most
               recent release of the NIST/EPA/MSDC mass spectral library
               shall be used  as the reference library.  The data system must
               be  capable of  flagging all data files that have been edited
               manually by laboratory personnel.

      3.7.6    Magnetic tape  storage device - must be capable of recording
               data  and must  be suitable for long-term, off-line storage.

      3.7.8    pH  paper - wide  range.

4.    Reagents

4.1   Reagent water - defined as  water in which an interferent is not
      observed at or above the CRQL of the parameters of interest.

      4.1.1    Reagent water  may be generated by passing tap water through a
               carbon filter  bed containing about 453 g of activated  carbon
               (Calgon Corp., Filtrasorb-300 or equivalent).

      4.1.2    A water purification system  (Millipore Super-Q or equivalent)
               may be used to generate reagent water.

      4.1.3    Reagent water  may also be prepared by boiling water for  15
               minutes.  Subsequently, while maintaining the temperature at
               90°C, bubble a contaminant-free  inert gas through the  water
               for one hour.  While still hot,  transfer the water co  a
               narrow-mouth screw-cap bottle and seal with a Teflon-lined
               septum and cap.

4.2   Sodium thiosulfate - (ACS)  granular.

4.3   Methanol - pesticide quality or equivalent.

5.    Standards

5.1   The Contractor must provide all standards to be  used with this
      contract.  These standards may be used only after they have been
      certified according to the procedure in Exhibit  E.  The Contractor must
      be able  to verify that the standards are certified.  Manufacturer's
      certificates  of analysis must be retained by the Contractor and
      presented upon request.

                                 D-19/VOA                       OLM01.3  2/91

-------
                                                              SECTION  IV
5.2   Stock Standard Solutions

      Stock standard solutions may be purchased or may be prepared  in
      methanol from pure standard materials.

      5.2.1   Prepare stock standard solutions by placing about 9.8 mL of
              methanol into a 10.0 mL ground-glass stoppered volumetric
              flask.  Allow the flask to stand, unstoppered, for about 10
              minutes, or until all alcohol wetted surfaces have dried.
              Weigh the flask to the nearest 0.1 mg.

      5.2.2   Add the assayed reference material as described below.

              5.2.2.1    If the compound is a liquid,  using a 100 uL
                         syringe, immediately add two  or more drops of
                         assayed reference material  to the  flask, then
                         reweigh.  The liquid must fall  directly into the
                         alcohol without contacting  the  neck of the flask.

              5.2.2.2    If the compound is a gas at room temperature,  fill
                         a 5 mL valved gas-tight syringe with the reference
                         standard to the 5.0 mL mark.  Lower the needle to  5
                         mm above the methanol meniscus.  Slowly introduce
                         the reference standard above  the surface of the
                         liquid.  The gas will rapidly dissolve in the
                         methanol.   This may also be accomplished by using  a
                         lecture bottle equipped with  a  Hamilton Lecture
                         Bottle Septum (#86600).  Attach Teflon tubing to
                         the side-arm relief valve and direct a gentle
                         stream of the reference standard into the methanol
                         meniscus.

      5.2.3   Reweigh, dilute to volume, stopper, then mix by inverting the
              flask several times.  For non-gaseous and compounds, calculate
              the concentration in micrograms per microliter from the net
              gain in weight.  When compound purity is assayed to be 97
              percent or greater, the weight may be used, without
              correction, to calculate the concentration of the stock
              standard.  If the compound purity is assayed to be less than
              97 percent, the weight must be corrected when calculating the
              concentration of the stock solution.   For gaseous compounds,
              calculate the concentration in micrograms  per microliter,
              using the Ideal Gas Law, taking into account the temperature
              and pressure conditions within the laboratory.

      5.2.4   Prepare fresh stock standards every two  months for gases or
              for reactive compounds such as styrene.   All other stock
              standards for non-gases/non-reactive purgeable compounds must
              be replaced after six months, or sooner  if standard has
              degraded or evaporated.
                                 D-20/VOA                       OLM01.5  4/91

-------
                                                              SECTION IV
5;3   Secondary Dilution Standards

      5.3.1   Using stock standard solutions, prepare secondary dilution
              standards in methanol that contain the compounds of interest,
              either singly or mixed together.  Secondary dilution standard
              solutions should be prepared at concentrations that can be
              easily diluted to prepare working standard solutions.

      5.3.2   Prepare fresh secondary dilution standards for gases and for
              reactive compounds such as styrene every month, or sooner, if
              standard has degraded or evaporated.  Secondary dilution
              standards for the other purgeable compounds must be replaced
              after six months, or sooner if standard has degraded or
              evaporated.

5.4   Working Standards

      5.4.1   Instrument Performance Check Solution - p-Bromofluorobenzene
              (BFB)

              Prepare a 25 ng/uL solution of BFB in methanol.  Prepare fresh
              BFB solution every six months, or sooner, if the solution has
              degraded or evaporated.  NOTE:  The 25 ng/uL concentration is
              used with a 2 uL injection volume.  The laboratory may prepare
              a 50 ng/uL solution of BFB if a 1 uL injection volume is used.

      5.4.2   Calibration Standard Solution

              Prepare the working calibration standard solution containing
              all of the purgeable target compounds in methanol.  The
              recommended concentration of the target compounds is 100
              ug/mL.  Prepare fresh working calibration standard solutions
              weekly, or sooner, if solutions have degraded or evaporated.

      5.4.3   Internal Standard Spiking Solution

              Prepare an internal standard spiking solution containing
              Bromochloromethane, Chlorobenzene-d5, and 1,4-Difluorobenzene
              in methanol at the concentration of 25.0 ug/mL for each
              internal standard.  Add 10 uL of this spiking solution into
              5.0 mL of sample or calibration standard for a concentration
              of 50 ug/L.  Prepare fresh spiking solution weekly, or sooner
              if the solution has degraded or evaporated.

      5.4.4   System Monitoring Compound (SMC) Spiking Solution

              Prepare a system monitoring compound spiking solution
              containing Toluene-dg, p-Bromofluorobenzene, and 1-2-
              Dichloroethane-d4 in methanol at a concentration of 25.0
              ug/mL.  Add 10.0 uL of this spiking solution into 5.0 mL of
              sample, for a concentration of 50 ug/L.  Prepare fresh spiking
              solution weekly,  or sooner if the solution has degraded or
              evaporated.


                                 D-21/VOA                      OLM01.5  4/91

-------
                                                              SECTION IV
      5.4.5   Volatile Matrix Standard Spiking Solution

              5.4.5.1    Prepare a spiking solution in methanol that
                         contains the following compounds at a concentration
                         of 25.0 ug/mL:   1,1-Dichloroethene,
                         Trichloroethene,  Chlorobenzene,  Toluene, and
                         Benzene.  Prepare fresh spiking solution weekly, or
                         sooner if the solution has degraded or evaporated.

              5.4.5.2    Matrix spikes are analyzed in duplicate; therefore,
                         add an aliquot of this solution to each of two
                         portions from one sample chosen for spiking.

5.5   Aqueous Calibration Standard Solutions

      5.5.1   Prepare five aqueous initial calibration standard solutions
              containing all of the purgeable target compounds and system
              monitoring compounds at the 10, 20, 50, 100, and 200 ug/L
              levels.  Note:  These are not the same levels as have been
              used in previous Statements of Work.  It is required that all
              three Xylene isomers (o-,  p-, and m-Xylene) be present in the
              calibration standards at concentrations of each isomer equal
              to that of the other target compounds (i.e., 10, 20, 50, 100,
              and 200 ug/L).  Similarly, the cis and trans isomers of 1,2-
              dichloroethene must both be present in the standards at
              concentrations of each isomer equal to that of the other
              target compounds.

      5.5.2   Aqueous calibration standards may be prepared in a volumetric
              flask or in the syringe used to inject the standard into the
              purging device.

              5.5.2.1    Volumetric flask - add an appropriate volume of
                         working calibration standard solution to an aliquot
                         of reagent water in a volumetric flask.   Use a
                         microsyringe and rapidly inject the alcohol
                         standard into the expanded area of the filled
                         volumetric flask.  Remove the needle as quickly as
                         possible after injection.  Bring to volume.  Mix by
                         inverting the flask three times  only.  Discard the
                         contents contained in the head of the flask.

              5.5.2.2    Syringe - remove  the plunger from a 5 mL "Luerlock"
                         syringe.   Pour  reagent water into the syringe
                         barrel  to just  short of overflowing.   Replace the
                         syringe plunger and compress the water.   Invert the
                         syringe,  open the syringe valve  and vent any
                         residual air.   Adjust the water  volume to 5.0 mL
                         minus the amount  of calibration  standard to be
                         added.   Withdraw the plunger slightly and add an
                         appropriate volume of working calibration standard
                         through the valve bore of the syringe.   Close the
                         valve and invert  three times.


                                 D-22/VOA                       OLM01.5  4/91

-------
                                                              SECTION IV


      5.5.3    The  50  ug/L aqueous  calibration standard solution is the
               continuing  calibration  standard.

      5.5.4    The  methanol purged  in  each of the aqueous calibration
               standards must not exceed 1% by volume.

5.6   Storage of Standards

      5.6.1    Store the stock standards in Teflon-sealed screw-cap bottles
               with zero headspace  at  -10°C to -20°C.  Protect  the standards
               from light.  Once one of the bottles containing  the stock
               standard solution has been opened, it may be used for no longer
               than one week.

      5.6.2    Store secondary dilution standards in Teflon-sealed screw-cap
               bottles with minimal headspace at -10°C to -20°C.  Protect the
               standards from light.   The secondary dilution standards must be
               checked frequently for  signs of degradation or evaporation,
               especially  just prior to preparing the working calibration
               standards from them.

      5.6.3    Aqueous standards may be stored for up to 24 hours if held in
               Teflon-sealed screw-cap vials with zero headspace at 4°C.
               Protect the standards from light. If not so stored, they must
               be discarded after one  hour unless they are set  up to be purged
               by an autosarapler.   When using an autosampler, the standards
               may  be  kept for up to 12 hours in purge tubes connected via the
               autosampler to the purge and trap device.

      5.6.4    Purgeable standards  must be stored separately from other
               standards.

6.     Instrument Operating Conditions

6.1   Purge and Trap Device

      The following are the recommended purge and trap analytical conditions
      except as stated below:

      Purge Conditions:

            Purge Gas:                     Helium or Nitrogen
            Purge Time:                     11.0 + 0.1 min
            Purge Flow Rate:                25-40 mL/min
            Purge Temperature:              Ambient (water or medium
                                           level soil),  required
                                           40°C (low level soil),
                                           required

      Desorb Conditions:

            Desorb Temperature:             180"C
            Desorb Flow Rate:               15 mL/min
            Desorb Time:                    4.0 + 0.1 min

                                 D-23/VOA                      OLM01.5  4/91

-------
      Trap Reconditioning Conditions:

            Reconditioning Temperature:     180°C
            Reconditioning Time:            7.0 rain + 0.1 min

      Before initial use, condition the  trap overnight at 180°C by
      backflushing with at least 20 mL/min flow of inert gas.   Vent the trap
      effluent to the room and not to the analytical column.   Prior to daily
      use, condition the trap by heating at 180°C for 10 minutes while
      backflushing.  The trap may be vented to the analytical  column during
      daily conditioning; however, the column must be run through the
      temperature program prior to analysis of samples.

6.2   Gas Chromatograph

      The following are the recommended GC analytical conditions:

      6.2.1   Packed Columns

              Carrier Gas:                      Helium
              Flow Rate:                        30 mL/min
              Initial Temperature:              45°C
              Initial Hold Time:                3 min

              Ramp Rate:                        8°C/min
              Final Temperature:                220°C
              Final Hold Time:                  15 min

              Transfer Line Temperature:        250-300°C

      6.2.2   Capillary Columns

              Carrier Gas:                      Helium
              Flow Rate:                        15 mL/min
              Initial Temperature:              10°C
              Initial Hold Time:                1.0 - 5.0 min  (±0.1 min)
              Ramp Rate:                        6°C/min
              Final Temperature:                160°C
              Final Hold Time:                  Until all target
                                                 compounds elute

      6.2.3   Optimize GC conditions for analyte separation and sensitivity.
              Once optimized, the same GC conditions must be used for  the
              analysis of all standards, samples, blanks, matrix spikes, and
              matrix spike duplicates.
                                 D-24/VOA                     OLM01.1   12/90

-------
6.3   Mass Spectrometer

      The following are the required mass spectrometer conditions:

      Electron Energy:                     70 Volts (nominal)
      Mass Range:                          35-300 amu
      Scan Time:                           To give at least 5 scans per
                                           peak, not to exceed 2 seconds per
                                           scan for capillary column

                                           To give at least 5 scans per
                                           peak, not to exceed 3 seconds per
                                           scan for packed column

6.4   The GC/MS system must be tuned to meet the manufacturer's
      specifications,  using a suitable calibrant such as FC-43 or
      perfluorokerosene (PFK).  The mass calibration and resolution of the
      GC/MS system are verified by the analysis of the instrument performance
      check solution (paragraph 5.4.1).

      6.4.1   Prior to the analyses of any samples, blanks, or calibration
              standards, the  Contractor must establish that the GC/MS system
              meets the mass  spectral  ion abundance criteria  for the
              instrument performance check solution containing p-
              bromofluorobenzene  (BFB).

      6.4.2   The analysis of the instrument performance check solution may
              be performed as follows:

              o  As  an injection of  up  to  50  ng of  BFB  into the GC/MS.

              o  By  adding  50  ng of  BFB to 5.0  ml  of  reagent water and
                 analyzing  the resulting solution  as  if  it  were an
                 environmental sample  (see section  8  below).

              BFB may not be  analyzed  simultaneously with  a calibration
              standard.

      6.4.3   The mass spectrum of BFB must be acquired in the following
              manner.   Three  scans  (the peak apex scan and the scans
              immediately preceding and following the apex) are acquired and
              averaged.  Background subtraction is required,  and must be
              accomplished using a single scan prior to the elution of BFB.
              NOTE:  All subsequent standards, samples, MS/MSD, and blanks
              associated with a BFB analysis must use identical mass
              spectrometer instrument  conditions.
                                 D-25/VOA                      OLM01.8   8/91

-------
      6.4.4   The analysis of the  instrument performance check solution must
              meet  the  ion abundance criteria given below.

                                TABLE 1
                    BFB  KEY IONS AND ION ABUNDANCE CRITERIA

              Mass      Ion Abundance Criteria

              50        8.0-40.0 percent of mass 95
              75        30.0-66.0 percent of mass 95
              95        base peak,  100 percent relative abundance
              96        5.0-9.0 percent of mass 95 (see note)
              173       less than 2.0 percent of mass  174
              174       50.0 -  120.0 percent of mass 95
              175       4.0 - 9.0 percent of mass 174
              176       93.0 -  101.0 percent of mass 174
              177       5.0 - 9.0 percent of mass 176

            Note:   All ion abundances must be normalized to m/z  95,  the
            nominal base peak,  even though the ion abundance of  m/z  174 may
            be up to 120 percent that of m/z 95.
      6.4.5   The criteria listed above are based on adherence to the
              acquisition specifications identified in paragraph 6.4.3, and
              were developed for the specific target compound list associated
              with this Statement of Work.  The criteria are based on
              performance characteristics of instruments currently utilized
              in routine support of Program activities.  These
              specifications, in conjunction with relative response factor
              criteria for 23 target compounds (see Table 2), are designed to
              control and monitor instrument performance associated with the
              requirements of this Statement of Work.

      6.4.6   The instrument performance check solution must be injected once
              at the beginning of each 12-hour period, during which samples
              or standards are to be analyzed.  The twelve (12) hour time
              period for GC/MS Instrument Performance Check  (BFB), standards
              calibration (initial or continuing calibration criteria) and
              method blank analysis begins at the moment of injection of the
              BFB analysis that the laboratory submits as documentation of a
              compliant instrument performance check.  The time period ends
              after twelve (12) hours has elapsed according to the system
              clock.

7.     Calibration

7.1   Prior to the analysis of samples and required blanks,  and after the
      instrument performance check solution criteria have been met,  each
      GC/MS system must be calibrated at a minimum of five concentrations to
      determine instrument sensitivity and the linearity of GC/MS response
      for the purgeable target compounds.
                                 D-26/VOA                             OLM01.0

-------
7.2   Assemble a purge and trap device that meets the specification in 3.6.
      Condition the trap overnight at 180°C in the purge  mode  with an inert
      gas flow of at least 20 mL/min.  Daily,  prior to use,  condition the
      traps for 10 minutes while backflushing  at 180°C with  the column at
      220°C.

7.3   Connect the purge and trap device to a gas c.hromatograph.   The gas
      chromatograph must be operated using temperature and flow rate
      parameters equivalent to those in 6.2.   Calibrate the  purge and
      trap-GC/MS system using the internal standard technique  (7.4).

7.4   Internal standard calibration procedure.   The three internal standards
      are Bromochloromethane,  1,4-Difluorobenzene, and Chlorobenzene-d5, at
      50 ug/L at time of purge.   Separate initial and continuing calibrations
      must be performed for water samples, and low level  soil  samples
      (unheated purge vs.  heated purge).   Extracts of medium level soil
      samples may be analyzed using the calibrations for  water samples.

      7.4.1   Prepare calibration standards at a minimum of five
              concentration levels for each target compound and system
              monitoring compound, as specified in 5.5.  Standards may be
              stored up to 24 hours, following the procedures in paragraph
              5.6.3.

      7.4.2   Prepare a spiking solution containing each of the internal
              standards using the procedures described in paragraph 5.4.3.

      7.4.3   Verify that the GC/MS system meets the instrument performance
              criteria in paragraph 6.4 by injecting BFB.  Analyze  each
              calibration standard, according to paragraph 7.1, adding  10 uL
              of internal standard spiking solution directly to the syringe.

              Tabulate the area response of the characteristic ions in  the
              extracted ion current profile (EICP) against concentration for
              each compound and internal standard and calculate relative
              response factors (RRF) for each compound as follows:

                                    Ax          cis
                        RRF   =     —    x     —
                                    Ais         Cx

              Where,

              Ax  =    Area of the  characteristic  ion (EICP) for  the compound
                       to  be measured (see Table  4)

              Ais =    Area of the  characteristic  ion (EICP) for  the
                       specific  internal  standard (see  Table 3)

              Cis =    Concentration  of the internal  standard

              Cx  =    Concentration  of the compound  to be measured
                                 D-27/VOA                            OLM01.0

-------
         Calculating the  relative  response factor of the Xylenes  and  the
        -cis  and trans  isomers  of  1,2-Dichloroethene requires  special
         attention.   On packed  columns, o-and p-Xylene  isomers coelute.
         On capillary columns,  the m- and p-Xylene  isomers coelute.
         Therefore,  when  calculating the relative response factor in  the
         equation above,  use  the area response  (Ax) and concentration
         (Cx)  of the peak that  represents the single isoraer on the GC
         column  used for  analysis.

         For  the cis and  trans  isomers  of 1,2-Dichloroethene which may
         coelute on  packed columns but  not on capillary columns,  both
         isomers must be  present in the standards.  If  the two isomers
         coelute,  use the area  of  the coeluting peak and the total
         concentration  of the two  isomers in the standard to determine
         the  relative response  factor.  If the  two  isomers do  not
         coelute,  sum the areas of the  two peaks and the concentrations
         of the  two  isomers in  the standard to  determine the relative
         response factor.

7.4.4    The  average relative response  factor (RRF) must be calculated
         for  all compounds.  Calculate  the % Relative Standard Deviation
         (%RSD)  of RRF  values over the  working  range of the curve.
                   %RSD = Standard deviation  x 100
                                mean
        Where,
                                        - *)'
        Standard Deviation  =
                                  n
                                       n-1
1/2
        Where,

        Xt = each individual value used to calculate the mean

        x  = the mean of n values

        n  = the total number of values

7.4.5   The response factors of the compounds listed below  (Table  2)
        must meet the minimum RRF criteria at each concentration level
        and maximum %RSD criteria for the initial calibration,  with
        allowance made for up to two volatile compounds.  However, the
        RRFs for those two compounds must be greater than or  equal to
        0.010, and the %RSD of those two compounds must be  less than or
        equal to 40.0% for the initial calibration to be acceptable.
                           D-28/VOA                       OLM01.2  1/91

-------
                                   TABLE 2
         RELATIVE RESPONSE FACTOR CRITERIA FOR INITIAL AND CONTINUING
                   CALIBRATION OF VOLATILE ORGANIC COMPOUNDS

Volatile                         Minimum            Maximum           Maximum
Compound	RRF	%RSD	%Diff

Bromomethane                       0.100             20.5              25.0
Vinyl chloride                     0.100             20.5              25.0
1,1-Dichloroethene                 0.100             20.5              25.0
1,1-Dichloroethane                 0.200             20.5              25.0
Chloroform                         0.200             20.5              25.0
1,2-Dichloroethane                 0.100             20.5              25.0
1,1,1-Trichloroethane              0.100             20.5              25.0
Carbon tetrachloride               0.100             20.5              25.0
Bromodichloromethane               0.200             20.5              25.0
cis-l,3-Dichloropropene            0.200             20.5              25.0
Trichloroethene                    0.300             20.5              25.0
Dibromochloromethane               0.100             20.5              25.0
1,1,2-Trichloroethane              0.100             20.5              25.0
Benzene                            0.500             20.5              25.0
trans-l,3-Dichloropropene          0.100             20.5              25.0
Bromoform                          0.100             20.5              25.0
Tetrachloroethene                  0.200             20.5              25.0
1,1,2,2-Tetrachloroethane          0.500             20.5              25.0
Toluene                            0.400             20.5              25.0
Chlorobenzene                      0.500             20.5              25.0
Ethylbenzene                       0.100             20.5              25.0
Styrene                            0.300             20.5              25.0
Xylenes (total)                    0.300             20.5              25.0
Bromofluorobenzene                 0.200             20.5              25.0
      7.4.6    The  following  compounds have no Maximum %RSD, or Maximum
               %Difference  criteria; however, these compounds must meet  a
               minimum RRF  criterion of 0.010:

               Acetone                           1,2-Dichloropropane
               2 -Butanone                        2 -Hexanone
               Carbon disulfide                  Methylene chloride
               Chloroethane                     4-Methyl- 2-pentanone
               Chloromethane                     Toluene-dg
               1,2-Dichloroethene  (total)        1,2-Dichloroethane-d4

      7.4.7    A check of the calibration  curve must be performed once every
               12 hours  (see  paragraph 6.4.6  for  the definition of the twelve
               hour time period).   Check the  relative response factors of
               those compounds for which RRF  values have been established.  If
               these criteria are  met, the relative response factors  for all
               compounds are  calculated and reported.  A percent difference of
               the  daily relative  response factor  (12 hour) compared  to  the
               average relative  response factor from the initial curve is

                                 D-29/VOA                      OLM01.2   1/91

-------
                                                              SECTION IV
               calculated.  Calculate the percent difference for each compound
               and  compare  with  the maximum percent difference criteria listed
               above.   For  negative percent difference values, the value must
               be greater than or equal to -25.0%, but less than-0%.  As with
               the  initial  calibration, up to two volatile compounds in Table
               2 may fail to meet the minimum RRF or maximum %D criteria, but
               the  RRFs of  those two compounds must be greater than or equal
               to 0.010, and the percent differences must be less than or
               equal to 40.0% for the continuing calibration to be acceptable.

      7.4.8    Internal standard responses and retention times in all
               standards must be evaluated during or immediately after data
               acquisition. If the retention time for any internal standard
               changes  by more than 0.50 minutes (30 seconds) from the latest
               daily (12 hour) calibration standard, the chromatographic
               system must  be inspected for malfunctions, and corrections made
               as required.  The extracted ion current profile (EICP) of the
               internal standards must be monitored and evaluated for each
               standard.  If the EICP area for any internal standard changes
               by more  than a factor of two (-50% to -4-100%), the mass
               spectrometric system must be inspected for malfunction, and
               corrections  made as appropriate.  When corrections are made,
               re-analysis  of samples analyzed while the system was
               malfunctioning is required.

7.5   Each GC/MS system must be calibrated upon award of the contract,
      whenever the Contractor takes corrective action which may change or
      affect the initial calibration criteria (i.e.,  i*on source cleaning or
      repair, column removal or replacement,  etc.),  or if the continuing
      calibration acceptance criteria have not been met'.

7.6   If time remains  in the 12 hour time period after meeting the acceptance
      criteria for the initial calibration,  samples may be analyzed.   It is
      not necessary to analyze a continuing calibration standard,  if the
      initial calibration meets the calibration acceptance criteria above.  A
      method blank is necessary.  Quantify all sample results against the
      initial calibration standard that is the same concentration as the
      continuing calibration standard (50 ug/L).

7.7   If time does not remain in the 12-hour period beginning with the
      injection of the instrument performance check solution,  a new injection
      of the instrument performance check solution must be made.  If the new
      injection meets  the ion abundance criteria for BFB,  then a continuing
      calibration standard may be injected.

7.8   The concentrations of volatile target compounds in the continuing
      calibration standard are given in paragraph 5.5.3.

7.9   The response factors for the continuing calibration standard must meet
      the criteria given in paragraph 7.4.5 prior to the analysis of any
      blanks or samples.
                                 D-30/VOA                       OLM01.3  2/91

-------
                                                              SECTION  IV
8.     Sample Analysis

8 .1   Water Samples

      8.1.1   All water samples must be allowed to warm to ambient
              temperature before analysis.

      8.1.2   Prior to the analysis of samples, establish the appropriate
              GC/MS operating conditions, as outlined in paragraphs 6-6.4.6,
              analyze the instrument performance check solution (6.4), and
              calibrate the GC/MS system according to paragraphs 7-7.7.3.

      8.1.3   If time remains in the 12-hour period (as described in
              paragraph 7.6), samples may be analyzed without analysis of a
              continuing calibration standard.

      8.1.4   If time does not remain in the 12-hour period since the
              injection of the instrument performance check solution, both
              the instrument performance check solution and the continuing
              calibration standard must be analyzed before sample analysis
              may begin (see paragraphs 7.7-7.9).

      8.1.5   Adjust the purge gas (helium) flow rate to 25-40 mL/min.
              Variations from this flow rate may be necessary to achieve
              better purging and collection efficiencies for some compounds,
              particularly Chloromethane and Bromoform.

      8.1.6   Remove the plunger from a 5 mL syringe and attach a closed
              syringe valve.  Open the sample or standard bottle which has
              been allowed to come to ambient temperature, and carefully pour
              the sample into the syringe barrel to just short of
              overflowing.  Replace the syringe plunger and compress the
              sample.  Open the syringe valve and vent any residual air while
              adjusting the sample volume to 5.0 mL.  This process of taking
              an aliquot destroys the validity of the sample for future
              analysis so, if there is only one VOA vial, the analyst should
              fill a second syringe at this time to protect against possible
              loss of sample integrity.  This second sample is maintained
              only until such time as the analyst has determined that the
              first sample has been analyzed properly.  Filling one 5 mL
              syringe would allow the use of only one syringe.  If an
              analysis is needed from the second 5 mL syringe, it must be
              performed within 24 hours. Care must also be taken to prevent
              air from leaking into the syringe.

      8.1.7   Once the sample aliquots have been taken from the VOA vial, the
              pH of the water sample must be determined.  The purpose of the
              pH determination is to ensure that all VOA samples were
              acidified in the field.  Test the pH by placing one or two
              drops of sample on the pH paper (do not add pH paper to the
              vial).  Record the pH of each sample, and report these data in
              the SDG Narrative, following the instructions in Exhibit B.  No
              pH adjustment is to be performed by the Contractor.


                                 D-31/VOA                      OLM01.2   1/91

-------
                                                        SECTION IV
8.1.8   The purgeable organics screening procedure (Section III), if
        used, will have shown the approximate concentrations of major
        sample components.  If a dilution of the sample was indicated,
        this dilution shall be made just prior to GC/MS analysis of the
        sample.  All steps in the dilution procedure must be performed
        without delays until the point at which the diluted sample is
        in a gas tight syringe.

        8.1.8.1    The following procedure will allow for dilutions
                   near the calculated dilution factor from the
                   screening procedure:

                   8.1.8.1.1    All dilutions are made in volumetric
                                flasks (10 mL to 100 mL).

                   8.1.8.1.2    Select the volumetric flask that will
                                allow for the necessary dilution.
                                Intermediate dilutions may be necessary
                                for extremely large dilutions.

                   8.1.8.1.3    Calculate the approximate volume of
                                reagent  water which will  be added to
                                the volumetric flask selected and add
                                slightly less than this quantity of
                                reagent  water to the flask.

                   8.1.8.1.4    Inject the proper aliquot from the
                                syringe  prepared in paragraph 8.1.6
                                into the volumetric flask.   Aliquots of
                                less than 1 mL increments are
                                prohibited.   Dilute the flask to the
                                mark with reagent water.   Cap the
                                flask, invert, and shake  three times.

                   8.1.8.1.5    Fill a 5 mL syringe with the diluted
                                sample as in paragraph 8.1.6.

                   8.1.8.1.6    If this  is an intermediate dilution,
                                use it and repeat the above procedure
                                to achieve larger dilutions.

8.1.9   Add 10.0 uL of the system monitoring compound  spiking  solution
        (paragraph 5.4.4) and 10.0 uL of the  internal  standard spiking
        solution (paragraph 5.4.3) through the valve bore  of the
        syringe, then close the valve.  The system monitoring  compounds
        and internal standards may be mixed and added  as a single
        spiking solution.  The addition of 10 uL of  the system
        monitoring compound spiking solution  to 5 mL of sample is
        equivalent to a concentration of 50 ug/L of  each system
        monitoring compound.

8.1.10  Attach the syringe-syringe valve assembly  to the syringe valve
        on the purging device.  Open  the syringe valves and  inject  the
        sample into the purging chamber.

                           D-32/VOA                             OLM01.0

-------
                                                        SECTION IV
8.1.11  Close both valves and purge the sample for 11.0 + 0.1 minutes
        at ambient temperature.

8.1.12  At the conclusion of the purge time, attach the trap to the
        chroraatograph, adjust the device to the desorb mode, and begin
        the gas chromatographic temperature program.  Concurrently,
        introduce the trapped materials to the gas chromatographic
        column by rapidly heating the trap to 180°C while backflushing
        the trap with an inert gas between 20 and 60 mL/min for four
        minutes.  If this rapid heating requirement cannot be met, the
        gas chromatographic column must be used as a secondary trap by
        cooling it to 30°C (or subambient, if problems persist) instead
        of the recommended initial temperature of 45°C.

8.1.13  While the trap is being desorbed into the gas chromatograph,
        empty the purging chamber.  Wash the chamber with a minimum of
        two 5 mL flushes of reagent water to avoid carryover of target
        compounds.

8.1.14  After desorbing the sample for four minutes, recondition the
        trap by returning the purge and trap device to the purge mode.
        Wait 15 seconds, then close the syringe valve on the purging
        device to begin gas flow through the trap.  The trap
        temperature should be maintained at 180°C.  Trap temperatures
        up to 220°C may be employed, however the higher temperature
        will shorten the useful life of the trap.  After approximately
        seven minutes, turn off the trap heater and open the syringe
        valve to stop the gas flow through the trap.  When cool, the
        trap is ready for the next sample.

8.1.15  Each analytical run must be checked also for saturation.  The
        level at which an individual compound will saturate the
        detection system is a function of the overall system
        sensitivity and the mass spectral characteristics of that
        compound.  The initial calibration requires that the system
        should not be saturated for high response compounds at 200 ug/L
        for VGA target compounds.  In addition, the system must not be
        saturated by the two Xylene isomers that coelute on the GC
        column used for analysis when the coeluting peak represents 400
        ug/L, or, for the two 1,2-Dichloroethene isomers that may
        coelute when the coeluting peak represents 400 ug/L.  Secondary
        ion quantitation is allowed only when there are sample matrix
        interferences with the primary ion.  If secondary ion
        quantitation is performed, document the reasons in the SDG
        Narrative.  When a sample is analyzed that has saturated ions
        from a compound, this analysis must be followed by the analysis
        of a reagent water blank.  If the blank is not free of
        interferences, the system must be decontaminated.  Sample
        analysis may not resume until a blank has been analyzed that
        demonstrates that the system is free of interferences.  Once
        the system is free of interferences, the sample that saturated
        the detection must be diluted and reanalyzed.
                           D-33/VOA                       OLM01.2   1/91

-------
                                                        SECTION IV
8.1.16   To  prepare  a matrix spike and matrix spike duplicate for water
         samples.,  add 10 uL of  the matrix spike solution (paragraph
         5.4.5)  to each of the   5 mL aliquots of the sample chosen for
         spiking.  Disregarding any dilutions, this is equivalent to a
         concentration of 50 ug/L of each matrix spike compound.  The
         frequency of MS/MSD analysis is given in paragraph 10.8.

8.1.17   A volatile  method blank must be analyzed at least once during
         every twelve hour time period, on each GC/MS system used for
         volatile  analysis (see paragraph 6.4.6 for the definition of
         the twelve  hour time period).

         8.1.17.1    For water samples,  a volatile method blank consists
                    of a 5 mL volume of reagent water (paragraph 4.1)
                    spiked with  the system monitoring compounds and
                    internal standards,  and carried through the
                    analytical  procedure.

         8.1.17.2   An acceptable volatile method blank for water
                    samples must contain less than or equal to five
                    times (5x)  the Contract Required Quantitation Limit
                    (CRQL,  see  Exhibit C)  of Methylene chloride,
                   Acetone,  and 2-Butanone, and less than or equal to
                    the CRQL of any other volatile target compound.

         8.1.17.3   All volatile analyses associated with a blank that
                   does not meet the requirements above, (i.e., a
                   contaminated blank)  must be repurged, reanalyzed,and
                   reported at no additional cost to the Agency.

         8.1.17.4   The volatile method blank must be analyzed after the
                   calibration standards,  to ensure that there is no
                   carryover of material  from the standards into
                   samples.

8.1.18   The  Contractor must demonstrate that there is no carryover  from
         a contaminated sample  before data from subsequent analyses may
         be  submitted.  After a sample that contains a target compound
         at  a level  exceeding the initial calibration range, the
         Contractor  must either:

         8.1.18.1   Analyze a method blank immediately after the
                   contaminated sample.  If an autosampler is used, a
                   method blank must also be analyzed using the same
                   purge inlet  that was used for the contaminated
                   sample.   The method blanks must meet the technical
                   acceptance criteria for blank analysis (see 8.1.17),
                   or

         8.1.18.2   Monitor the sample analyzed immediately after the
                   contaminated sample for all compounds that were in
                   the contaminated sample and that exceeded the limits
                   above.   The maximum contamination criteria are as
                   follows:   the sample must not contain a

                          D-34/VOA                      OLM01.2   1/91

-------
                                                              SECTION IV
                         concentration above the CRQL for the target
                         compounds that exceeded the limits in the
                         contaminated sample.  If an autosampler is used, the
                         next sample analyzed using the same purge inlet that
                         was used for the contaminated sample also must meet
                         the maximum contamination criteria.

8.2   Soil/Sediment Samples

      Two approaches may be taken to determine whether the low level or
      medium level method must be followed.

      o Assume the sample  is low level and analyze a 5 g sample.

      o Use the X factor calculated from the optional hexadecane screen
        (Section III, paragraph 6.2.1.3) to determine the appropriate method
        for analysis.

      If peaks are saturated from the analysis of a 5 g sample,  a smaller
      sample size must be analyzed to prevent saturation.   However, the
      smallest sample size permitted is 1 g.   If smaller than 1 g sample size
      is needed to prevent saturation,  the medium level method musjt be used.

      8.2.1   Low Level Soil Method

              The low level soil method is based on a heated purge of a
              soil/sediment sample mixed with reagent water containing the
              system monitoring compounds and the internal standards.
              Analyze all method blanks and standards under the same
              conditions as the samples.

              Use 5 grams of sample, or use the X Factor to determine the
              sample size for purging.

              o  If  the  X  Factor  is  0  (no  peaks  noted on the hexadecane
                 screen),  analyze  a  5 g sample.

              o  If  the  X  Factor  is  between  0 and 1.0,  analyze a minimum of a
                 1 g sample.

              o  If  the  X  Factor  is  >1,  use  the  medium soil  method.


              8.2.1.1    The GC/MS system should be set up as in paragraphs
                         7-7.7.3.  This should be done prior to the
                         preparation of the sample to avoid loss of volatiles
                         from standards and sample.  A heated purge
                         calibration curve must be prepared and used for the
                         quantitation of all samples analyzed with the
                         low-level method.  Follow the initial and daily
                         calibration instructions (7.4 and 7.7), but increase
                         the purge temperature to 40°C.
                                 D-35/VOA                       OLM01.2   1/91

-------
                                                SECTION  IV
8.2.1.2    To prepare the reagent water containing the system
           monitoring compounds and the internal standards,
           remove the plunger from a 5 mL "Luerlock"  type
           syringe equipped with a syringe valve and  fill until
           overflowing with reagent water.  Replace the plunger
           and compress the water to vent trapped air.  Adjust
           the volume to 5.0 mL.   Add 10 uL of the system
           monitoring compound spiking solution and 10 uL of
           the internal standard solution to the syringe
           through the valve.

8.2.1.3    The sample (for volatile organics)  consists of the
           entire contents of the sample container.  Do not
           discard any supernatant liquids.  Mix the  contents
           of the sample container with a narrow metal spatula.
           Weigh the amount determined in paragraph 8.2.1 into
           a tared purge device.   Use a top loading balance.
           Note and record the actual weight to the nearest  0.1
           g.

8.2.1.4    Add the spiked reagent water to the purge  device  and
           connect the device to the purge and trap system.
           NOTE:   Prior to the attachment of the purge device,
           the steps in paragraphs 8.2.1.2 and 8.2.1.3 must  be
           performed rapidly to avoid loss of  volatile
           organics.  These steps must be performed in a
           laboratory free of solvent fumes.

8.2.1.5    Immediately after weighing the sample,  weigh 5-10 g
           of the sediment into a tared crucible.   Determine
           the percent moisture by drying overnight at 105°C.
           Allow to cool in a desiccator before weighing.
           Concentrations of individual analytes will be
           reported relative to the dry weight of sediment.


           g of wet sample-g of dry sample
           g of wet sample                  x 100 -  % moisture

8.2.1.6    Heat the sample to 40°C + 1°C and purge the sample
           for 11.0 + 0.1 minutes.

8.2.1.7    Proceed with the analysis as outlined in paragraphs
           8.1.11 - 8.1.14.  Requirements for  dilution of
           samples are given in paragraphs 8.2 and 10.7.
                   D-36/VOA                      OLM01.2  1/91

-------
                                               SECTION IV
8.2.1.8    To prepare  a matrix  spike  and matrix spike duplicate
           for low level soils/sediment, add  10 uL of the
           matrix spike solution  (5.4.4) to the 5 mL of water
           added to each of the two aliquots  of the soil from
           the sample  chosen for  spiking (paragraph 8.2.1.2).
           The concentration for  a 5  g  sample would be
           equivalent  to 50 ug/kg of  each matrix spike
           compound.   The frequency of  MS/MSB analysis is  given
           in paragraph 10.8.

8.2.1.9    A volatile  method blank must be analyzed at least
           once during every twelve hour time period, on each
           GC/MS system used for  volatile analysis (see
           paragraph 6.4.6 for  the definition of the twelve
           hour time period).

           8.2.1.9.1     For  low level soil/sediment samples, a
                        volatile  method blank consists of  a 5 g
                        of a purified solid matrix added to
                        reagent water,  spiked with the system
                        monitoring compounds  and internal
                        standards, and  carried through the
                        analytical procedure.

           8.2.1.9.2     An acceptable volatile method blank for
                        low level soil  samples must contain
                        less than or  equal to five times (5x)
                        the  Contract Required Quantitation
                        Limit (CRQL,  see Exhibit C) of
                        Methylene chloride, Acetone, and 2-
                        Butanone, and less than or equal to the
                        CRQL of any other volatile target
                        compound.

           8.2.1.9.3     All  volatile  analyses associated with a
                        blank that does not meet the
                        requirements  above, (i.e., a
                        contaminated blank) must be repurged,
                        reanalyzed,and  reported at no
                        additional cost to the Agency.

           8.2.1.9.4     The volatile  method blank must be
                        analyzed  after  the calibration
                        standards, to ensure  that there is no
                        carryover of  material from the
                        standards into  samples.

8.2.1.10   The Contractor must  demonstrate  that  there  is no
           carryover from a contaminated sample before data
           from subsequent analyses may be  submitted.  After  a
           sample that contains a target compound at a level
           exceeding the initial  calibration  range,  the
           Contractor  must either:
                   D-37/VOA                       OLM01.2  1/91

-------
                                                        SECTION IV
                   ft. 2.1.10.1   Analyze a method blank immediately
                                after the contaminated sample.   If an
                                autosampler is used,  a method blank
                                must also be analyzed using the  same
                                purge inlet that was  used for the
                                contaminated sample.   The method blanks
                                must meet the technical acceptance
                                criteria for blank analysis (see
                                8.2.1.9), or

                   8.2.1.10.2   Monitor the sample analyzed immediately
                                after the contaminated sample for  all
                                compounds that were in the contaminated
                                sample and that exceeded the limits
                                above.   The maximum contamination
                                criteria are as follows:  the sample
                                must not contain a concentration above
                                the CRQL for the target compounds  that
                                exceeded the limits in the contaminated
                                sample.  If an autosampler is used, the
                                next sample analyzed using the same
                                purge inlet that was  used for the
                                contaminated sample also must meet the
                                maximum contamination criteria.

8.2.2   Medium Level Soil Method

        The medium level soil method is based on extracting  the soil/
        sediment sample with methanol.  An aliquot of the methanol
        extract is added to reagent water containing the system
        monitoring compounds and the internal standards.  The reagent
        water containing the methanol extract is purged at ambient
        temperature.  All samples with an X Factor >1.0 must be
        analyzed by the medium level method.  If saturated peaks
        occurred, or would occur, when a 1 g sample was analyzed,   the
        medium level method must be used.

        8.2.2.1    The GC/MS system should be set up as in paragraphs
                   7-7.7.3. This should be done prior to the addition
                   of the methanol extract to reagent water.  Because
                   the methanol extract and reagent water mixture  is
                   purged at ambient temperature, the instrument
                   performance check, initial calibration, and
                   continuing calibration for water samples may be used
                   for analyses of medium soil sample extracts.

        8.2.2.2    The sample (for volatile organics) consists of the
                   entire contents of the sample container.  Do not
                   discard any supernatant liquids.  Mix the contents
                   of the sample container with a narrow metal spatula.
                   Weigh 4 g (wet weight) into a tared 15 mL vial.  Use
                   a top loading balance.  Note and record the actual
                   weight to the nearest 0.1 g.  Determine the percent
                   moisture as in paragraph 8.2.1.5.

                           D-38/VOA                       OLM01.2   1/91

-------
                                                              SECTION IV
               8.2.2.3    Quickly add 9.0 mL of methanol to the vial.  Then
                         add 1.0 mL of the system monitoring compound spiking
                         solution to the vial.  Cap and shake for 2 minutes.
                         NOTE:  The steps in paragraphs 8.2.2.2 and 8.2.2.3
                         must be performed rapidly to avoid loss of volatile
                         organics.  These steps must be performed in a
                         laboratory free of solvent fumes.

               8.2.2.4    Using a disposable pipette, transfer approximately 1
                         mL of extract into a GC vial for storage.  The
                         remainder may be discarded.  Transfer approximately
                         1 mL of the reagent methanol to a GC vial for use as
                         the method blank for each Case, SDG, or day on which
                         medium soil sample extractions are performed,
                         whichever is most frequent.  These extracts may be
                         stored in the dark at 4°C (±2°C) prior to analysis.

               8.2.2.5    The following table can be used to determine the
                         volume of methanol extract to add to the 5 mL of
                         reagent water for analysis.  If the hexadecane
                         screen procedure was followed, use the X factor
                         (Option B) or the estimated concentration (Option A)
                         to determine the appropriate volume.  Otherwise,
                         estimate the concentration range of the sample from
                         the low level analysis to determine the appropriate
                         volume.  If the sample was submitted as a medium
                         level sample, start with 100 uL.

                         All dilutions must keep the response of the major
                         constituents (previously saturated peaks) in the
                         upper half of linear range of the curve.
                                 Estimated                Take this Volume of
        X Factor           Concentration Range             Methanol Extract
ug/kg
0.25
0.5
2.5
12.5
5.0
- 10.0
- 50.0
- 250
500 -
1000 -
5000 -
25,000 -
10,000
20,000
100,000
500,000
uL
100
50
10
100 of 1/50 dilution3
Calculate appropriate dilution factor for concentrations exceeding the table.

    Actual  concentration ranges  could be 10  to  20  times  higher than this  if
    the  compounds are halogenated and the estimates  are  from GC/FID.
2
    The  volume  of methanol added to the  5 mL of water being purged should be
    kept constant.   Therefore, add to the 5  mL  syringe whatever volume of
    methanol  is necessary to maintain a  volume  of  100 uL added to the syringe.
Q
    Dilute  an aliquot of the methanol extract and  then take 100 uL for
    analysis.
                                 D-39/VOA                             OLM01.0

-------
                                                SECTION  IV
8.2.2.6    Remove the plunger from a 5 mL "Luerlock"  type
           syringe equipped with a syringe valve and  fill until
           overflowing with reagent water.  Replace the plunger
           and compress the water to vent trapped air.   Adjust
           the volume to 4.9 mL.  Pull the plunger back to 5 mL
           to allow volume for the addition of sample and
           standards.  Add 10 uL of the internal standard
           solution.  Also add the volume of methanol extract
           determined in paragraph 8.2.2.5 and a volume of
           clean methanol to total 100 uL (excluding  methanol
           in standards).

8.2.2.7    Attach the syringe-syringe valve assembly  to the
           syringe valve on the purging device.   Open the
           syringe valve and inject the water/methanol sample
           into the purging chamber.

8.2.2.8    Proceed with the analysis as outlined in paragraph
           8.   Analyze all method blanks on the  same  instrument
           as the samples.  Requirements for dilution of
           samples are given in paragraphs 8.2 and 10.7.

8.2.2.9    To prepare a matrix spike and matrix  spike duplicate
           for the medium level soil/sediment samples,  add 8.0
           mL of methanol, 1.0 mL of the system  monitoring
           compound spiking solution,  and 1.0 mL of matrix
           spike solution (paragraph 5.4.4) as in paragraph
           8.2.2.3,  to each of the two aliquots  of the soil
           sample chosen for spiking.   This results in a 6,200
           ug/kg concentration of each matrix spike compound
           when added to a 4 g sample.   Add a 100 uL  aliquot of
           this extract to 5 mL of water for purging  (as per
           paragraph 8.2.2.6).   The frequency of MS/MSD
           analysis is given in paragraph 10.8.

8.2.2.10.  A volatile method blank must be analyzed at least
           once during every twelve hour time period,  on each
           GC/MS system used for volatile analysis (see
           paragraph 6.4.6 for the definition of the  twelve
           hour time period).

           8.2.2.10.1   For medium level soil/sediment  samples,
                        a  volatile method blank  consists  of a 4
                        g  of a  purified solid matrix  spiked
                        with the system monitoring compounds,
                        extracted with  methanol,  and  carried
                        through the analytical procedure.
                  D-40/VOA                       OLM01.3  2/91

-------
                                               SECTION IV
         ••.. 8.2.2.10.2   An  acceptable volatile method blank  for
                       medium level soil/sediment samples must
                       contain less than or equal five  times
                       (5x) the Contract Required Quantitation
                       Limit (CRQL, see Exhibit C) of
                       Methylene chloride, Acetone, and 2-
                       Butanone, and less than or equal ot  the
                       CRQL of any other volatile target
                       compound.

           8.2.2.10.3   All volatile analyses associated with  a
                       blank that does not meet the
                       requirements above, (i.e. a
                       contaminated blank) must be repurged,
                       reanalyzed,and reported at no
                       additional cost to the Agency.

           8.2.2.10.4   The volatile method blank must be
                       analyzed after the calibration
                       standards, to ensure that there  is no
                       carryover of material from the
                       standards into samples.

8.2.2.11   The Contractor must demonstrate that there  is no
           carryover from a contaminated sample before data
           from subsequent  analyses may be submitted.  After a
           sample  that  contains a target compound at a level
           exceeding the initial calibration range, the
           Contractor must  either:

           8.2.2.11.1   Analyze a method blank immediately
                       after the contaminated sample.   If an
                       autosampler is used, a method blank
                       must also be analyzed using the  same
                       purge inlet that was used for the
                       contaminated sample.  The method blanks
                       must meet the technical acceptance
                       criteria for blank analysis (see
                       8.2.2.10), or

           8.2.2.11.2   Monitor the sample analyzed immediately
                       after the contaminated sample for all
                       compounds that were in the contaminated
                       sample and that exceeded the limits
                       above.  The maximum contamination
                       criteria are as follows:  the sample
                       must not contain a concentration above
                       the CRQL for the target compounds that
                       exceeded the limits in the contaminated
                       sample.  If an autosampler is used,  the
                       next sample analyzed using the same
                       purge inlet that was used for the
                       contaminated sample also must meet the
                       maximum contamination criteria.

                  D-41/VOA                      OLM01.2  1/91

-------
                                                              SECTION  IV
9.     Qualitative Analysis

9.1   The compounds listed in the Target Compound List  (TCL),  Exhibit  C,
      shall be identified by an analyst competent in the  interpretation of
      mass spectra (see Exhibit A,  Section III)  by comparison  of the sample
      mass spectrum to the mass spectrum of a standard  of the  suspected
      compound.   Two criteria must be satisfied  to verify the
      identifications: (1) elution of the sample component at  the same GC
      relative retention time as the standard component,  and  (2)
      correspondence of the sample component and standard component mass
      spectra.

      9.1.1   For establishing correspondence of the GC relative retention
              time  (RRT),  the sample component RRT must compare within +0.06
              RRT units of the RRT of the standard component.   For reference,
              the standard must be run in the same 12-hour time period as the
              sample.  If samples are analyzed during the same 12-hour time
              period as the initial calibration standards, use the RRT values
              from  the 50 ug/L standard.  If coelution of interfering
              components prohibits accurate assignment of the sample
              component RRT from the total ion chromatogram,  the RRT should
              be assigned by using extracted ion current profiles for ions
              unique to the component of interest.

      9.1.2   For comparison of standard and sample component mass spectra,
              mass spectra obtained on the Contractor's GC/MS are required.
              Once obtained, these standard spectra may be used for
              identification purposes, only if the Contractor's GC/MS meets
              the daily instrument performance requirements for BFB.   These
              standard spectra may be obtained from the run used to obtain
              reference RRTs.

      9.1.3   The requirements for qualitative verification by comparison of
              mass spectra are as follows:

              9.1.3.1    All ions present in the standard mass spectra at a
                         relative intensity greater than 10%  (most abundant
                         ion in the spectrum equals 100%) must be present in
                         the sample spectrum.

              9.1.3.2    The relative intensities of ions specified in
                         paragraph 9.1.3.1 must  agree within + 20% between
                         the standard and sample spectra.  (Example:   For an
                         ion with an abundance of 50% in the  standard
                         spectra,  the corresponding sample abundance  must be
                         between 30 and 70 percent).
                                 D-42/VOA                       OLM01.3   2/91

-------
                                                              SECTION IV
               9.1.3.3    Ions greater than 10% in the sample spectrum but not
                         present in the standard spectrum must be considered
                         and accounted for by the analyst making the
                         comparison.  In Task III, the verification process
                         should favor false positives.  All compounds meeting
                         the identification criteria must be reported with
                         their spectra.  For all compounds below the CRQL
                         report the actual value followed by a "J", e.g.,
                         113 J."

      9.1.4    If a compound cannot be verified by all of the criteria in
               paragraph 9.1.3.3, but in the technical judgement of the mass
               spectral interpretation specialist, the identification is
               correct, then the Contractor shall report that identification
               and proceed with quantification in paragraph 10.

9.2   A library search shall be executed for non-target  sample components for
      the purpose of tentative identification.   For  this purpose,  the most
      recent release of the NIST/EPA/MSDC mass spectral  library,  shall be
      used.  Computer generated library search routines  must not use
      normalization routines that would misrepresent the library or unknown
      spectra when compared to each other.

      9.2.1    Up to  10 organic compounds of greatest apparent concentration
               not listed in Exhibit C for the purgeable organic fraction,
               excluding the system monitoring compounds, shall be tentatively
               identified via a forward search of the NIST/EPA/MSDC Library
               (substances with responses less than 10% of the internal
               standard are not required to be searched in this fashion).
               Only after visual comparison of sample spectra with  the nearest
               library searches will the mass spectral interpretation
               specialist assign a tentative identification.  Computer
               generated library search routines .must not use normalization
               routines that would misrepresent the library or unknown spectra
               when compared to each other.

      9.2.2    Guidelines for making tentative identification:

               9.2.2.1    Relative intensities of major ions in the reference
                         spectrum (ions greater than 10% of the most abundant
                         ion) should be present in the sample spectrum.

               9.2.2.2    The relative intensities of the major ions should
                         agree within + 20%.  (Example:   For an ion with an
                         abundance of 50 percent of the standard spectra, the
                         corresponding sample ion abundance must be between
                         30 and 70 percent.)

               9.2.2.3    Molecular ions present in reference spectrum should
                         be present in sample spectrum.
                                 D-43/VOA                       OLM01.3   2/91

-------
                                                              SECTION IV
               9.2.2.4    Ions present in the sample spectrum but not in the
                         reference spectrum should be reviewed for possible
                         background contamination or presence of co-eluting
                         compounds.

               9.2.2.5    Ions present in the reference spectrum but not in
                         the sample spectrum should be reviewed for possible
                         subtraction from the sample spectrum because of
                         background contamination or co-eluting compounds.
                         Data system library reduction programs can sometimes
                         create these discrepancies.

      9.2.3    If, in the technical judgement of the mass spectral
               interpretation specialist, no valid tentative identification
               can be made, the compound should be reported as unknown.  The
               mass spectral specialist should give additional classification
               of the unknown compound, if possible  (i.e., unknown aromatic,
               unknown hydrocarbon, unknown acid type, unknown chlorinated
               compound).  If probable molecular weights can be distinguished,
               include them.

10.   Quantitative Analysis

10.1  Target components identified shall be quantified by the internal
      standard method.   The internal standard used shall be  that which is
      assigned in Table 5 of this Section.   The  EICP area of the
      characteristic ions of analytes listed in  Tables 3 and 4 in this
      Section are used.

      In all instances  where the data system report has been edited, or where
      manual integration or quantitation has been performed,  the GC/MS
      operator must  identify such edits  or  manual procedures by initializing
      and dating the changes made to the report.

10.2  Internal standard responses and retention  times in all standards must
      be evaluated during or immediately after data acquisition.   If the
      retention time for any internal standard changes by more than 30
      seconds from the  latest daily  (12  hour) calibration standard,  the
      chromatographic system must be inspected for malfunctions,  and
      corrections made  as required.   For samples analyzed during the same 12-
      hour time period  as the initial calibration standards,  compare the
      internal standard responses and retention  times against the 50 ug/L
      calibration standard.   The extracted  ion current profile (EICP) of the
      internal standards must be monitored  and evaluated for each sample,
      blank, matrix spike,  and matrix spike duplicate.  If the EICP area for
      any internal standard changes  by more than a factor of two (-50% to
      +100%), the mass  spectrometric system must be inspected for malfunction
      and corrections made  as appropriate.   When corrections are made,
      reanalysis of samples analyzed while  the system was malfunctioning is
      necessary.
                                 D-44/VOA                      OLM01.2   1/91

-------
                                                              SECTION IV
      10.2.1  If after re-analysis, the EICP areas for all internal standards
              are inside the contract limits (-50% to +100%), then the
              problem with the first analysis is considered to have been
              within the control of the laboratory.  Therefore, submit only
              data from the analysis with EICPs within the contract limits.
              This is considered the initial analysis and must be reported as
              such on all data deliverables.

      10.2.2  If the re-analysis of the sample does not solve the problem,
              i.e., the EICP areas  are outside the contract limits for both
              analyses, then submit the EICP data and sample data from both
              analyses.  Distinguish'between the initial analysis and the
              re-analysis on all data deliverables, using the sample suffixes
              specified in Exhibit B.  Document in the SDG Narrative all
              inspection and corrective actions taken.

10.3  The relative response factor (RRF)  from the  continuing calibration
      standard is  used to calculate the concentration in the sample.   Use the
      relative response factor as determined  in paragraph 7.4.3 and the
      equations  below.   When target compounds are  below contract required
      quantitation limits (CRQL),  but the spectra  meet the identification
      criteria,  report the concentration with a "J."   For example,  if CRQL is
      10 ug/L and  concentration of 3 ug/L is  calculated,  report as  "3J."
                                               (Ax)(Is)(Df)
              Concentration         ug/L =
                                              (Ais)(RRF)(V0)

              Where,

              Ax  =     Area of the characteristic ion (EICP)  for the compound
                        to be measured (see Table 4)
              A^s =     Area of the characteristic ion (EICP)  for the specific
                        internal standard (see Table  3)
              Is  =     Amount of internal standard added in nanograms (ng)
              RRF =     Relative response factor from the ambient temperature
                        purge of the calibration standard.
              Vo  =     Volume of water purged in milliliters (mL)

              Df  =     Dilution factor.   The dilution factor for analysis of
                        water samples for volatiles by this method is defined as
                        the ratio of the number of milliliters (mL) of water
                        purged (i.e., V0 above) to the number of mL of the
                        original water sample used for purging.   For example, if
                        2.5 mL of sample is diluted to 5.0 mL with reagent water
                        and purged, Df = 5.0 mL/2.5 mL = 2.0.   If no dilution is
                        performed, Df = 1.0.
                                 D-45/VOA                             OLM01.0

-------
                                                SECTION IV


Low Soil

Concentration                      (AX)(IS)
(Dry weight basis)    ug/Kg - (Ais)(BRF)(Wa)(D)

Where ,

Ax, Is, Ais are as given for water.


RRF =     Relative response factor from the  heated purge of the
          calibration standard.

D   =     100 - %  moisture
                100

Ws  -     Weight of sample  added to the purge tube,  in grams (g)


Medium Soil

Concentration                  (Ax) (Is) (Vt) (1000) (Df )
(Dry weight basis)    ug/Kg -   (Al8)(RRF)(Va)(Ws)(D)

Where,

Ax, A^g, Is are as given for water above.
RRF  =    Relative response factor from the ambient temperature
          purge of the calibration standard.
Vt  -     Total volume of the methanol extract in milliliters
          (mL) .   NOTE:  This volume is typically 10.0 mL,  even
          though only 1.0 mL is  transferred to the vial in
          paragraph 8.2.2.4.

Va  =•     Volume of the aliquot  of the methanol extract in
          microliters (uL) added to reagent water for purging

Ws  =     Weight of soil extracted, in grams (g)

D   =     100 -  % moisture
               100

Df  =     Dilution factor.  The  dilution factor for analysis of
          soil/sediment samples  for volatiles by the medium level
          method is defined as the ratio of the number of
          microliters (uL) of methanol added to the reagent water
          for purging i.e., Va above,  to the number of microliters
          of the methanol extract of the sample contained in that
          volume Va.   The dilution factor is equal to 1.0 in all
          cases other than those requiring dilution of the
          methanol extract.  Dilution of the extract is required
          when the "X" factor (paragraph 8.2.2.5) is > 12.5.

          The factor of 1,000 in the numerator converts the value
          of Vt from mL to uL.

                   D-46/VOA                       OLM01.2   1/91

-------
                                                              SECTION IV
10.4  An estimated concentration for non-target components tentatively
      identified shall be determined by the internal standard .method.   For
      quantification, the nearest internal standard free of interferences
      shall be used.

      The formula for calculating concentrations is the same as in paragraph
      10.3.  Total area counts (or peak heights) from the total ion
      chromatograms are to be used for both the compound to be measured and
      the internal standard.  A relative response factor (RRF) of one  (1) is
      to be assumed.  The resulting concentration shall be qualified as "J"
      (estimated, due to lack of a compound-specific response factor), and
      "N" (presumptive evidence of presence),  indicating the quantitative and
      qualitative uncertainties associated with this non-target component.
      An estimated concentration should be calculated for all tentatively
      identified compounds as well as those identified as unknowns.

10.5  Xylenes (o-,  m-,  and p- isomers) are to be reported as Xylenes (total).
      Because the o- and p-Xylene isomers coelute on packed columns, and the
      m- and p-Xylene isomers coelute on capillary columns, special attention
      must be given to the quantitation of the Xylenes.  The relative
      response factor (RRF) determined in paragraph 7.4 is based on the peak
      that represents the single isomer on the GC column used (m-Xylene on
      packed columns, o-Xylene on capillary columns).  In quantitating sample
      concentrations, use the areas on both peaks and the RRF from paragraph
      7.4.   The areas of the two peaks may be summmed,  and the concentration
      determined, or the concentration represented by each of the two  peaks
      may be determined separately, and then summed.  It is required that all
      three Xylene isomers be present in the initial and continuing
      calibration standards.

10.6  The CLS and trans steriosomers of 1,2-Dichloroethene are to be reported
      as 1,2-Dichloroethene (total).  If the two isomers coelute on the GC
      column used for analysis, use the area of the coeluting peaks and the
      RRF determined in 7.4 to determine the concentration.  If the isomers
      do not coelute, use the single RRF values determined in 7.4 to
      determine the concentration.   The areas of the two peaks may be  summed
      and the concentration determined, or the concentration represented by
      each of the two peaks may be determined separately, and then summed.
      It is required that both the cis and trans isomers of the 1,2-
      Dichloroethene be present in the initial and continuing calibration
      standards.

10.7  If the on-column concentration of any compound in any sample exceeds
      the initial calibration range, a new aliquot of that sample must be
      diluted and purged.  Guidance in performing dilutions, and exceptions
      to this requirement are given below.

      10.7.1  Use the results  of the  original  analysis  to  determine the
              approximate dilution  factor  required  to  get  the  largest analyte
              peak within the  initial  calibration range.

      10.7.2  The dilution  factor chosen should keep the response  of  the
              largest analyte peak  for a target compound in the  upper half of
              the initial calibration  range of the  instrument.

                                 D-47/VOA                       OLM01.2  1/91

-------
                                                              SECTION IV
      10.7.3  Do not-submit data for more than two analyses, i.e., the
              original  sample and one dilution, or, if the volatile screening
              procedure was employed, from the most concentrated dilution
              analyzed  and one further dilution.

      10.7.4  Do not dilute MS/MSD samples to get either spiked or non-spiked
              analytes  within calibration range, if the sample from which the
              MS/MSD aliquots were taken contains high levels of the spiked
              analytes, calculate the concentration and recovery of the
              analytes  from the undiluted analysis, and note the problem in
              the SDG Narrative.

      10.7.5  For total Xylenes, where three isomers are quantified as two
              peaks, the calibration of each peak, should be considered
              separately, i.e., a diluted analysis is not required for total
              Xylenes unless the concentration of the peak representing the
              single isomer exceeds 200 ug/L or the peak representing the two
              coeluting isomers on the GC column exceeds 400 ug/L.
              Similarly, if the cis and trans isomers of 1,2-Dichloroethene
              coelute,  a diluted analysis is not required unless the
              concentration of the coeluting peak exceeds 400 ug/L.  If the
              two isomers do not coelute, a diluted analysis is not required
              unless the concentration of either peak exceeds 200 ug/L.

10.8  Calculate the recovery of each  system  monitoring compound in all
      samples, blanks,  matrix spikes,  and matrix spike duplicates.   Determine
      if the recovery is within limits (see  Table 6),  and report on
      appropriate form.

      10.8.1  Calculate the concentrations of the system monitoring compounds
              using the same equations as used for target compounds.
              Calculate the recovery of each system monitoring compound as
              follows:

              %Recovery =  Concentration (or amount) found    x 100
                           Concentration (or amount) spiked

      10.8.2  If the recovery of any one system monitoring compound is not
              within limits, the following are required:

              o   Check  to  be  sure  that  there  are no  errors  in calculations,
                  formulation of the  system  monitoring compound spiking
                  solutions,  and internal standards.   Also check instrument
                  performance.

              o   Reanalyze the  sample if none  of the  above  steps reveal a
                  problem.

              o   If an  undiluted analysis with acceptable monitoring
                  compound  recoveries  is  being  submitted,  do not reanalyze
                  diluted samples if  the  system monitoring compound
                  recoveries  are  outside  the limits.
                                 D-48/VOA                      OLM01.2   1/91

-------
                                                              SECTION IV
               o   .Never reanalyze the matrix spike or matrix spike duplicate
                  (MS/MSD), even if the system monitoring.compound recoveries
                  are outside the limits.
               o   If the sample associated with the matrix spike and matrix
                  spike duplicate does not meet specifications, it should be
                  reanalyzed only if the MS/MSD system monitoring compound
                  recoveries are within the limits.  If the sample and
                  associated MS/MSD show the same pattern (i.e., outside the
                  limits), then the sample does not require reanalysis and a
                  reanalysis must not be submitted.  Document in the
                  narrative the similarity in recoveries of the system
                  monitoring compounds in the sample and associated MS/MSD.

      10.8.3   If  the reanalysis of  the sample  solves the problem,  then  the
               problem was within the  laboratory's  control.  Therefore,  submit
               only  data from  the analysis with system monitoring compound
               recoveries within the limits.  This  shall be considered the
               initial analysis and  shall be reported as such on all data
               deliverables.

      10.8.4   If  the reanalysis of  the sample  does not solve the problem
               (i.e., the system monitoring compound recoveries  are outside
               the limits for both analyses), then  submit the data  from  both
               analyses.  Distinguish between the initial analysis  and the
               reanalysis on all data deliverables, using the sample suffixes
               supplied in Exhibit B.

      10.8.5   For medium level soil analyses,  involving methanol extraction,
               the treatment of system monitoring compound recoveries  is
               similar to that for semivolatile surrogate recoveries.  If any
               system monitoring compound recovery  is outside the limits,
               reanalyze the methanol extract first, to determine if the
               problem was with the  analysis.   If reanalysis of  the extract
               does  not solve  the problem, then reextract the medium soil
               sample and analyze the  second extract.  Follow paragraphs
               10.8.3 and 10.8.4 when determining which analyses to submit.

      10.8.6   If  the recovery of any one system monitoring compound in  a
               method blank is outside the limits,  then the method  and all
               associated samples must be reanalyzed at no additional cost  to
               the Agency.

10.9  A matrix spike and matrix spike duplicate must be performed for each
      group of samples of a similar matrix,  for the following,  whichever is
      most frequent:

      o   Each Case  of field samples received,  OR

      o   Each 20  field samples in a Case, OR

      o   Each group of field  samples  of  a similar  concentration level  (soils
          only), OR
                                 D-49/VOA                             OLM01.0

-------
                                                              SECTION IV
      o   Each 14 calendar day period (.7  calendar day period  for  14 -day  data
          turnaround contracts)  during which  field  samples  in a Case were
          received (said period beginning with  the  receipt  of the first  sample
          in that Sample Delivery Group) .

      10.9.1  Calculate  the concentrations of  the matrix spike compounds
              using the  same  equations as used for target  compounds.
              Calculate  the recovery of  each matrix spike  compound as
              follows :

                                       SSR -  SR
              Matrix Spike Recovery - -  x 100
              Where,

              SSR  - Spiked sample result
              SR   - Sample result
              SA   - Spike added


      10.9.2  Calculate the relative percent difference (RPD) of the
              recoveries of each compound in the matrix spike and matrix
              spike duplicate as follows:

              RPD -    |MSR - MSDR|    x 100
                     (1/2)(MSR+MSDR)

              Where,

              MSR - Matrix Spike Recovery

              MSDR = Matrix Spike Duplicate Recovery

              The vertical bars in the formula above indicate the absolute
              value of the difference, hence RPD is always expressed as a
              posit ive value .

      10.9.3  The limits for matrix spike compound recovery and RPD are given
              in Table 7.  As these limits are only advisory, no further
              action by the laboratory is required, however, frequent
              failures to meet the limits for recovery or RPD warrant
              investigation by the laboratory, and may result in questions
              from the Agency.

10.10 Determine  the  concentrations of any target compounds detected in the
      volatile method blank,  using the equations in paragraph 10.3.   The
      method blank must contain less  than or equal to the Contract Required
      Quantitation Limit (CRQL)  of the volatile target compounds in Exhibit
      C,  except  Methylene chloride, Acetone,  and 2-Butanone,  which must be
      less than  or equal to five times (5x)  the CRQL.   For soil/sediment
      method blanks,  CRQL value must  be adjusted for percent moisture (see
      Exhibit B) .
                                 D-50/VOA                       OLM01.2   1/91

-------
                                                        SECTION IV
If a laboratory method blank exceeds these criteria,  the Contractor
must consider the analytical system to be out of control.   The source
of the contamination must be investigated and appropriate corrective
measures MUST be taken and documented before further sample analysis
proceeds.  All samples processed with a method blank that is out of
control (i.e., contaminated) MUST be reextracted/repurged and
reanalyzed at no additional cost to the Agency.  The Laboratory
Manager, or his designee, must address problems and solutions in the
SDG Narrative (Exhibit B).
                           D-51/VOA                             OLM01.0

-------
                                                              SECTION IV
                                   TABLE 3
                           CHARACTERISTIC IONS FOR
                       SYSTEM MONITORING COMPOUNDS AND
              INTERNAL STANDARDS FOR VOLATILE ORGANIC COMPOUNDS
Compound	Primary Ion	Secondary lon(s)

SYSTEM MONITORING COMPOUNDS

4-Bromofluorobenzene                   95                     174, 176
l,2-Dichloroethane-d-4                 65                        102
Toluene-d-g                            98                      70, 100

INTERNAL STANDARDS

Bromochloromethane                    128                   49,  130, 51
1,4-Difluorobenzene                   114                      63,  88
Chlorobenzene-d-s                      117                      82, 119
                                 D-52/VOA                            OLM01.0

-------
                                                              SECTION IV
                                    TABLE  4
               CHARACTERISTIC  IONS  FOR VOLATILE TARGET COMPOUNDS
Analvte
Primary Ion*
          Secondary
            lonCs)
Chloromethane
Bromomethane
Vinyl chloride
Chloroethane
Methylene chloride
Acetone
Carbon disulfide
1,1-Dichloroethene
1,1-Dichloroethane
1,2-Dichloroethene
Chloroform
1,2-Dichloroethane
2-Butanone
1,1,1-Trichloroethane
Carbon tetrachloride
Bromodichloromethane
1,1,2,2-Tetrachloroethane
1,2 -Dichloropropane
trans-1,3-Dichloropropene
Trichloroethene
Dibromochloromethane
1,1,2-Trichloroethane
Benzene
cis-1,3-Dichloropropene
Bromoforra
2-Hexanone
4-Me thy1- 2 -pentanone
Tetrachloroethene
Toluene
Chlorobenzene
Ethyl benzene
Styrene
Total Xylenes
    50
    94
    62
    64
    84
    43
    76
    96
    63
    96
    83
    62
    43**
    97
   117
    83
    83
    63
    75
   130
   129
    97
    78
    75
   173
    43
    43
   164
    91
   112
   106
   104
   106
             52
             96
             64
             66
         49, 51, 86
             58
             78
           61,  98
     65, 83, 85, 98, 100
           61,  98
             85
        64,  100, 98
             57
        99,  117, 119
          119,  121
             85
      85,  131,  133, 166
           65,  114
             77
         95, 97, 132
          208,  206
     83, 85, 99, 132, 134

             77
171,  175,  250,  252, 254, 256
         58, 57, 100
           58,  100
       129,  131, 166
             92
            114
             91
           78,  103
             91
*   The  primary ion should be used unless interferences are present,  in which
    case,  a secondary ion may be used.

**_ m/z  43 is used for quantitation of 2-Butanone,  but m/z 72 must be present
    for  positive identification.
                                 D-53/VOA
                                       OLM01.0

-------
                                                              SECTION IV
                                   TABLE  5
        VOLATILE  INTERNAL STANDARDS WITH CORRESPONDING TARGET COMPOUNDS
          AND  SYSTEM MONITORING  COMPOUNDS ASSIGNED FOR QUANTITATION
Bromochlororaethane
1,4-Difluorobenzene
Chlorobenzene-d5
Ch1oromethane
Bromomethane
Vinyl Chloride
Chloroethane
Methylene Chloride
Acetone
Carbon Disulfide
1,1-Dichloroethene
1,1-Dichloroethane
1,2 -Dichloroethene(tot.)
Chloroform
1,2-Dichloroethane
2-Butanone
1,2-Dichloroethane-d4
  (smc)
1,1,1-Trichloroethane
Carbon Tetrachloride
Bromodichloromethane
1,2-Dichloropropane
trans -1,3-Dichloropropene
Trichloroethene
Dibromochloromethane
1,1,2-Trichloroethane
Benzene
cis-1,3-Dichloropropene
Bromoform
2-Hexanone
4-Methyl- 2 -Pentanone
Tetrachloroethene
1,1,2,2-Tetrachloroethane
Toluene
Chlorobenzene
Ethylbenzene
Styrene
Xylene (total)
Bromofluorobenzene (smc)
Toluene-d8 (smc)
(smc) - system monitoring compound
                                 D-54/VOA
                                            OLM01.0

-------
                                                              SECTION IV
                                   TABLE 6
                  SYSTEM MONITORING COMPOUND RECOVERY LIMITS
                               %Recovery           %Recovery
Compound	Water	Soil

Toluene-d8                      88-110              84-138
Bromofluorobenzene              86-115              59-113
l,2-Dichloroethane-d4           76-114              70-121
                                   TABLE 7
                          MATRIX SPIKE RECOVERY AND
                      RELATIVE PERCENT DIFFERENCE LIMITS
                               %Recovery     RPD       %Recovery   RPD
Compound	Water	Water	Soil	Soil

1,1-Dichloroethane              61-145      14         59-172     22
Trichloroethene          .       71-120      14         62-137     24
Benzene                         76-127      11         66-142     21
Toluene                         76-125      13         59-139     21
Chlorobenzene                   75-130      13         60-133     21
                                 D-55/VOA                             OLM01.0

-------
     EXHIBIT D
ANALYTICAL METHODS
 FOR  SEMIVOLATILES
    D-l/SV                              OLM01.0

-------
                              Table of Contents


                                                                 Page

SECTION I -   INTRODUCTION  	  D-3/SV

SECTION II -  SAMPLE PREPARATION AND STORAGE 	  D-5/SV

              PART A -  SAMPLE STORAGE AND HOLDING TIMES 	  D-6/SV

              PART B -  SAMPLE PREPARATION FOR EXTRACTABLE
                        SEMIVOLATILES (BNA) IN WATER 	  D-7/SV

              PART C -  SAMPLE PREPARATION FOR EXTRACTABLE
                        SEMIVOLATILES (BNA) IN
                        SOIL/SEDIMENT 	  D-12/SV

                        1.  Medium  Level  Preparation  for
                           Screening and Analysis  of
                           Semivolatiles (BNA) 	  D-13/SV

                        2.  Low  Level  Preparation  for
                           Screening  and Analysis  of
                           Semivolatiles (BNA) 	  D-17/SV

SECTION III - SCREENING OF SEMIVOLATILE ORGANIC EXTRACTS 	  D-32/SV

SECTION IV  - GC/MS ANALYSIS  OF SEMIVOLATILES 	  D-36/SV
                                  D-2/SV                              OLM01.0

-------
                                  SECTION I
                                 INTRODUCTION
The analytical methods that follow are designed to analyze water, soil and
sediment from hazardous waste sites for the organic compounds on the Target
Compound List (TCL, see Exhibit C).  The methods are based on EPA Method 625
(Bases/Neutrals and Acids).

The methods are divided into the following sections:  sample preparation,
screening, and analysis.  Sample preparation covers sample extraction and
cleanup techniques.  As described in the screening section, a portion of the
extracts may be screened on a gas chromatograph with appropriate detectors to
determine the concentration level of organics.   The analysis section contains
the GC/MS analytical methods for organics.
                                  D-3/SV                              OLM01.0

-------
                                                               SECTION I
1.     Method for the Determination of Extractable Semivolatile Organic
      Compounds.

1.1   Scope and Application

      This method covers the determination of a number of organic compounds
      that are partitioned into an organic solvent and are amenable to gas
      chromatography.  These target compounds and the contract required
      quantitation limits are listed in Exhibit C.

      Problems have been associated with the following compounds analyzed by
      this method:

          o  Dichlorobenzidine and 4-Chloroaniline can be subject to
             oxidative losses during solvent concentration.

          o  Hexachlorocyclopentadiene is subject to thermal decomposition in
             the inlet of the gas chromatograph, chemical reactions in
             acetone solution, and photochemical decomposition.
          o  N-Nitrosodiphenylamine decomposes in the gas chromatograph inlet
             forming diphenylamine and, consequently, may be detected as
             diphenylamine.
1.2   The method involves solvent extraction of the matrix sample,
      characterization to determine the appropriate analytical protocol to be
      used, and GC/MS analysis to determine semivolatile organic compounds
      present in the sample.
                                  D-4/SV                             OLM01.0

-------
          SECTION II
SAMPLE PREPARATION AND STORAGE
          D-5/SV                              OLM01.0

-------
                                                              SECTION II


PART A - SAMPLE STORAGE AND HOLDING TIMES

1.    Procedures for Sample Storage

      The samples must be protected from light and refrigerated at 4°C (±2°C)
      from the time of receipt until 60 days after delivery of a complete
      reconciled data package to the Agency.  After 60 days the samples may
      be disposed of in a manner that complies with all applicable
      regulations.

      Samples must be stored in an atmosphere demonstrated to be free of all
      potential contaminants.

      Samples, sample extracts, and standards must be stored separately.

2.    Procedure for.Sample Extract Storage

      Sample extracts must be protected from light and stored at 4°C (±2°C)
      until 365 days after delivery of a complete data package to the Agency.

      Sample extracts must be stored in an atmosphere demonstrated to be free
      of all potential contaminants.

      Samples, sample extracts, and standards must be stored separately.

3.    Contract Required Holding Times

      Extraction of water samples by continuous liquid-liquid procedures
      shall be started within 5 days of VTSR (Validated Time of Sample
      Receipt).   Extraction of soil/sediment samples by sonication procedures
      shall be completed within 10 days of VTSR.   NOTE:  Separatory funnel
      extraction procedures are not permitted.

      Extracts of either water or soil/sediment samples must be analyzed
      within 40 days following extraction.
                                  D-6/SV                        OLM01.3   2/91

-------
                                                              SECTION II


PART B - SAMPLE PREPARATION FOR EXTRACTABLE SEMIVOLATILES  (BNA)  IN WATER

1.     Summary of Sample Preparation Method

1.1   A one liter aliquot of sample is acidified to pH 2 and extracted with
      methylene chloride using a continuous liquid-liquid extractor.
      Separatory funnel extraction is NOT permitted.   The  methylene chloride
      extract is dried and concentrated to a volume of 1.0 mL.

2.     Interferences

      Contaminants in solvents, reagents, glassware,  and other sample
      processing hardware may cause method interferences such as discrete
      artifacts and/or elevated baselines in the total ion current profiles
      (TICPs).   All of these materials routinely must be demonstrated to be
      free from interferences under the conditions of the  analysis by running
      laboratory method blanks.  Matrix interferences may be caused by
      contaminants that are coextracted from the sample.  The extent of
      matrix interferences will vary considerably from source to source.

3 .     Apparatus and Materials

      Brand names, suppliers, and part numbers are for illustrative purposes
      only.  No endorsement is implied.  Equivalent performance  may be
      achieved using apparatus and materials other than those specified here,
      but demonstration of equivalent performance meeting the requirements of
      this SOW .is the responsibility of the Contractor.

3.1   Glassware (brand names and catalog numbers are included for
      illustration purposes only).

      3.1.1   Continuous liquid-liquid extractors  - equipped with Teflon or
              glass connnecting  joints and stopcocks requiring no lubrication
              (Hershberg-Wolf Extractor, Ace Glass Company, Vineland, NJ, P/N
              6841-10 or equivalent).

      3.1.2   Drying column  - 19 mm ID chromatographic  column with coarse
              frit  (substitution of a  small pad of Pyrex glass wool  for the
              frit will prevent  cross  contamination of  sample extracts).

      3.1.3   Concentrator tube  - Kuderna-Danish, 10 mL, graduated (Kontes
              K-570050-1025  or equivalent).  Calibration must be checked at
              the volumes employed in  the test.  Ground-glass stoppers  are
              used to prevent evaporation of extracts.

      3.1.4   Evaporative flask  - Kuderna-Danish, 500 mL (Kontes K-570001-
              0500 or equivalent).  Attach to concentrator  tube with springs.

      3.1.5   Snyder column  - Kuderna-Danish, three-ball macro  (Kontes
              K-503000-0121  or equivalent).

      3.1.6   Snyder column  - Kuderna-Danish, two-ball  micro (Kontes K569001-
              0219 or equivalent).


                                  D-7/SV                        OLM01.2   1/91

-------
                                                              SECTION II


      3.1.7   Vials  -.amber glass, 2 .mL capacity with Teflon-lined screw cap.

      3.1.8   Syringes  - 0.2 mL, 0.5 mL, and 5 mL volumes.

3.2   Silicon carbide boiling chips - approximately 10/40 mesh.   Heat to
      400°C for 30 minutes or Soxhlet extract with methylene chloride.

3.3   Water bath - heated, with concentric ring cover,  capable of temperature
      control (+ 2°C).  The bath should be used in a hood.

3.4   Balance - analytical, capable of accurately weighing + 0.0001 g.

3.5   Nitrogen evaporation device equipped with a water bath that can be
      maintained at 35-40°C.  The N-Evap by Organomation Associates,  Inc.,
      South Berlin, MA (or equivalent), is suitable.

4.    Reagents

4.1   Reagent water - defined as water in which an interferent is not
      observed at or above the CRQL of each parameter of interest.

4.2   Sodium thiosulfate - (ACS) granular.

4.3   Sulfuric acid solution (1+1) -  slowly add 50 mL of H2S04 (sp gr 1.84)
      to 50 mL of reagent water.

4.4   Acetone, methanol, methylene chloride - pesticide residue analysis
      grade or equivalent.

4.5   Sodium sulfate - (ACS) powdered, anhydrous.   Purify by heating at 400°C
      for four hours in a shallow tray, cool in a desiccator and store in a
      glass bottle (Baker anhydrous powder, catalog #73898,  or equivalent).

4.6   Surrogate standard spiking solution.

      4.6.1   Surrogate standards  are  added  to all  samples  and calibration
              solutions;  the compounds specified for this purpose are
              Phenol-d5,  2,4,6-Tribromophenol, 2-Fluorophenol,
              Nitrobenzene-d5,  Terphenyl-d]^, 2-Fluorobiphenyl,  2-
              Chlorophenol-d4,  and 1,2-Dichlorobenzene-d4.  Additional
              surrogates may be added  at the laboratory's discretion.

      4.6.2   Prepare a surrogate  standard spiking  solution that  contains
              Nitrobenzene-d5,  Terphenyl-d^4, 2-Fluorobiphenyl,  and  1,2-
              Dichlorobenzene-d4 at  a  concentration of  100  ug/mL; Phenol-d5,
              2,4,6-Tribromophenol,  2-Fluorophenol, and 2-Chlorophenol-d4  at
              a concentration  of 150 ug/mL.  Store  the  spiking solutions at
              4°C  (±2°C)  in Teflon-sealed  containers.   The  solutions should
              be checked  frequently  for stability.  These solutions  must be
              replaced  after twelve  months,  or sooner  if comparison  with
              quality control  check  samples  indicates  a problem.
                                  D-8/SV                        OLM01.2   1/91

-------
                                                              SECTION II
4.7.   BNA Matrix standard spiking solution -  the matrix spike  solution
      consists of the following:

            Bases/Neutrals                 Acids

            1,2,4-Trichlorobenzene       Pentachlorophenol
            Acenaphthene                 Phenol
            2,4-Dinitrotoluene           2-Chlorophenol
            Pyrene                       4-Chloro-3-methylphenol
            N-Nitroso-di-n-propylamine   4-Nitrophenol
            1,4-Dichlorobenzene

      Prepare a spiking solution that contains each of the base/neutral
      compounds above at 100 ug/1.0 mL in methanol and the acid compounds at
      150 ug/1.0 ml in methanol.   Store the spiking solutions  at 4°C (±2°C)
      in Teflon-sealed containers.  The solutions should be checked
      frequently for stability.  These solutions must be replaced after
      twelve months, or sooner if comparison with quality control check
      samples indicates a problem.

5.     Water Sample Extraction

5.1   Continuous liquid-liquid extraction is  used to extract the samples.

      5.1.1   Add methylene chloride to the bottom of the extractor and  fill
              it to a depth of at least 1 inch above the bottom side arm.

      5.1.2   Using a 1 liter graduated cylinder, measure out a 1.0 liter
              sample aliquot.  Transfer the 1 liter sample aliquot to  the
              continuous extractor.  Pipet 0.5 mL of surrogate standard
              spiking solution into the sample and mix well.  Check the  pH  of
              the sample with wide range pH paper and adjust the pH to 2.0
              with 1:1 H2S04.

      5.1.3   Following the procedures in 5.1.1 and 5.1.2 above, prepare two
              additional 1.0 Liter aliquots of the sample chosen for spiking.
              Add 0.5 mL of the BNA Matrix Spiking Solution to each of the
              additional aliquots.  The frequency of MS/MSD analysis is  given
              in Section IV, paragraph 8.6.

      5.1.4   Add 500 mL of methylene chloride to the distilling flask.  Add
              sufficient reagent water to ensure proper operation.  Extract
              for 18 hours.  Allow to cool, then detach the distilling flask,
              and label the flask.

      5.1.5   Prepare a method blank with each group of water samples
              extracted.  For semivolatile analyses, a method blank for  water
              samples consists of a 1 L volume of reagent water  (see
              paragraph 4.1), spiked with the surrogates and carried through
              the entire analytical procedure.  The frequency of method  blank
              analysis is given in Section IV, paragraph 8.7
                                  D-9/SV                      OLM01.1   12/90

-------
                                                              SECTION  II
5.2   Concentrating the Extracts

      5.2.1   Assemble a Kuderna-Danish (K-D) concentrator by attaching a 10
              mL concentrator tube to a 500 mL evaporative flask.  Other
              concentration devices or techniques may be used in place of the
              K-D, if equivalency is demonstrated .for all the semivolatile
              target compounds listed in Exhibit C.

      5.2.2   Transfer the extract by pouring the extract through a drying
              column containing about 10 cm of anhydrous granular sodium
              sulfate, and collect the extract in a K-D concentrator.  Rinse
              the Erlenmeyer flask and column with 20 to 30 mL of methylene
              chloride to complete the quantitative transfer.

      5.2.3   Add one or two clean boiling chips and attach a three-ball
              Snyder column to the evaporative flask.  Pre-wet the Snyder
              column by adding about 1 mL methylene chloride to the top of
              the column.  Place the K-D apparatus on a hot water bath (60°C
              to 80°C) so that the concentrator tube is partially immersed in
              the hot water, and the entire lower rounded surface of the
              flask is bathed with hot vapor.  Adjust the vertical position
              of the apparatus and the water temperature as required to
              complete the concentration in 10 to 15 minutes.  At the proper
              rate of distillation, the balls of the column will chatter
              actively, but the chambers will not flood with condensed
              solvent.  When the apparent volume of liquid reaches 1 mL,
              remove the K-D apparatus from the water bath and allow it to
              drain and cool for at least 10 minutes.  Remove the Snyder
              column and rinse the flask and its lower joint into the
              concentrator tube with 1 to 2 mL of methylene chloride.  A 5 mL
              syringe is recommended for this operation.

      5.2.4   Two different concentration techniques are permitted to obtain
              the final 1.0 mL volume:  micro Snyder column and nitrogen
              evaporation techniques.

              5.2.4.1    Micro Snyder  Column Technique

                         Add another one or two clean boiling chips to the
                         concentrator  tube and attach a two-ball micro Snyder
                         column.  Pre-wet the Snyder column by adding about
                         0.5 mL of methylene chloride to the top of the
                         column.  Place the K-D apparatus on a hot water bath
                         (60°C to 80°C) so that the concentrator tube is
                         partially immersed in the hot water.  Adjust the
                         vertical position of the apparatus and the water
                         temperature as required to complete the
                         concentration in 5 to 10 minutes.   At the proper
                         rate of distillation, the balls of the column will
                         chatter actively,  but the chambers will not flood
                         with condensed solvent.   When the apparent volume of
                         liquid reaches about 0.5 mL, remove the K-D
                         apparatus from the water bath and allow it to drain
                         for at least  10 minutes while cooling.   Remove the

                                 D-10/SV                       OLM01.2   1/91

-------
                                                              SECTION II
                         Snyder column and rinse its flask and its lower
                         joint into the concentrator tube with 0.2 mL of
                         methylene chloride.  Adjust the final volume to 1.0
                         mL with methylene chloride.  Transfer the extract to
                         a Teflon-sealed screw-cap bottle, label the bottle,
                         and store at 4°C (±2°C)..

               5^2.4.2    Nitrogen Evaporation Technique (taken from ASTM
                         Method D3086)

                         Place the concentrator tube with an open micro
                         Snyder attached in a warm water bath (30°C to 35°C)
                         and evaporate the solvent volume to just below 1 mL
                         by blowing a gentle stream of clean, dry nitrogen
                         (filtered through a column of activated carbon)
                         above the extract.  CAUTION:  Gas lines from the gas
                         source to the evaporation apparatus must be
                         stainless steel, copper, or Teflon tubing.  The
                         internal wall of the concentrator tube must be
                         rinsed down several times with methylene chloride
                         during the operation and the final volume brought to
                         1.0 mL with methylene chloride.  During evaporation,
                         the tube solvent level must be kept below the water
                         level of the bath.  The extract must never be
                         allowed to become dry.  Transfer the extract to a
                         Teflon-sealed screw-cap bottle, label the bottle and
                         store at 4°C (±2°C).

6.     The sample extracts are ready for GC/MS analysis.   Proceed to Section
      IV, GC/MS Analysis of Semivolatiles.   If high concentrations are
      suspected (e.g.,  highly co.lored extracts),  the optional GC/FID screen
      in Section III is recommended.
                                 D-ll/SV                       OLM01.2   1/91

-------
                                                              SECTION II
PART C - .SAMPLE PREPARATION FOR EXTRACTABLE SEMIVOLATILES (BNA) IN
SOIL/SEDIMENT

It is mandatory that all soil/sediment samples be characterized as to
concentration level so that the appropriate analytical protocol is chosen to
ensure proper quantitation limits for the sample.  Note that the terms "low
level" and "medium level" are not used here as a judgement of degree of
contamination but rather as a description of the concentration ranges that
are encompassed by the "low" and "medium" level procedures.

The laboratory is at liberty to determine the method of characterization.
The following two screening methods may be used for soil/sediment sample
characterization:

      o   Screen an aliquot from the  "low level"  30  g  extract or  an aliquot
          from the "medium level"  1 g extract.
      o   Screen using either GC/FID  or  GC/MS  as  the screening instrument.

The concentration ranges covered by these two procedures may be considered to
be approximately 330 ug/kg - 10,000 ug/kg for the low level analysis and
>10,000 ug/kg for medium level analysis for semivolatile extractables.

Screen from the Medium Level Method

Take 5.0 mL from the 10.0 mL total extract and concentrate to 1.0 mL and
screen. If the sample concentration is >10,000 ug/kg,  proceed with GC/MS
analysis of the organics.  If the sample concentration is <10,000 ug/kg,
discard the medium level extract and follow the low level method.

Screen from the Low Level Method

Take 5.0 mL from the 300 mL (approximate) total extract from the 30 g sample
and concentrate to 1.0 mL and screen.  If the original sample concentration
is >10,000 ug/kg, discard the 30 g extract and follow the medium level
methods for organics, using medium level surrogates.  If the sample
concentration is <10,000 ug/kg, proceed with concentration and the remainder
of the low level method.

Mandatory GPC Clean Up

Regardless of the concentration level, all soil/sediment sample extracts must
be subjected to clean up by Gel Permeation Chromatography (GPC).  Because the
effectiveness of GPC can be adversely affected by the amount of material
loaded onto the GPC column, it may be advisable to screen the sample extracts
described here prior to employing GPC.
                                  D-12/SV                              OLM01.0

-------
                                                              SECTION II


1.     Medi-um Level Preparation for Screening and Analysis of Semivolatiles

1.1   Scope and Application

      This procedure is designed for the preparation of soil/sediment samples
      which may contain organic chemicals at a level greater than 10,000
      ug/kg.

      1.1.1.  The extracts and sample aliquots prepared using this method are
              screened by GC/MS or FID, using capillary columns for
              semivolatile priority pollutants, and related organic
              chemicals.  The results of these screens will determine whether
              sufficient quantities of pollutants are present to warrant
              analysis by the medium protocol.

      1.1.2   If the screenings indicate no detectable pollutants at the
              lower limits of quantitation, the sample should be prepared by
              the low level protocol in Section II, Part C, beginning at
              paragraph 2.

1.2   Summary of Method

      1.2.1.  Approximately 1 g portions of soil/sediment are transferred to
              vials and extracted with methylene chloride.  The methylene
              chloride extract is screened for extractable organics by  GC/FID
              or GC/MS.

      1.2.2   If organic compounds are detected by the screen,, the methylene
              chloride extract is subjected to GPC clean up and analyzed by
              GC/MS for extractable organics.

      1.2.3   If no organic compounds are detected by the medium level
              screen, then a low level sample preparation is required.

1.3   Interferences

      1.3.1.  Method interferences may be caused by contaminants in solvents,
              reagents, glassware, and other sample processing hardware that
              lead to discrete artifacts and/or elevated baselines in the
              total ion current profiles.  All of these materials routinely
              must be demonstrated to be free from interferences under  the
              conditions of the analysis by running laboratory reagent
              blanks. Matrix interferences may be caused by contaminants  that
              are coextracted from the sample.  The extent of matrix
              interferences will vary considerably from source to source.

1.4   Limitations

      1.4.1.  The procedure is designed to allow quantitation limits for
              screening purposes as low as 10,000 ug/kg for extractable
              organics.  For analysis purposes, the quantitation limits are
              10,000 ug/kg for extractable organics.  If peaks are present
              based on the GC/FID screen, the sample is determined to require
              a medium level analysis by GC/MS.  Some samples may contain

                                 D-13/SV                       OLM01.2  1/91

-------
                                                              SECTION  II
              high concentrations of chemicals that interfere with the
              analysis of other components at lower levels;  the quantitation
              limits in those cases may be significantly higher.

      1.4.2   These extraction and preparation procedures were developed for
              rapid and safe handling of high concentration hazardous waste
              samples.  The design of the methods thus does not stress
              efficient recoveries or low limits of quantitation of all
              components.  Rather, the procedures were designed to screen, at
              moderate recovery and sufficient sensitivity,  a broad spectrum
              of organic chemicals.  The results of the analyses thus may
              reflect only a minimum of the amount actually present in some
              samples.
1.5   Reagents
      1.5.1.   Sodium Sulfate - anhydrous powdered reagent grade, heated at
              400°C for four hours, cooled in a desiccator, and stored in a
              glass bottle (Baker anhydrous powder, catalog # 73898 or
              equivalent).

      1.5.2   Acetone, Methanol, Methylene chloride - pesticide residue
              analysis grade or equivalent.

      1.5.3   Base/Neutral and Acid Surrogate Spiking Solution

              Surrogate standards are added to all samples and calibration
              solutions.  The compounds specified are Phenol-d5,
              2,4,6-Tribromophenol, 2-Fluorophenol, Nitrobenzene -d$,
              Terphenyl-d]^, 2-Fluorobiphenyl, 2-Chlorophenol-d4, and 1,2-
              Dichlorobenzene-d4.  Prepare a surrogate standard spiking
              solution that contains Nitrobenzene-d5, Terphenyl-d^,
              2-Fluorobiphenyl, and l,2-Dichlorobenzene-d4 at a concentration
              of 100 ug/mL; Phenol -6.5, 2,4,6-Tribromophenol, 2-Fluorophenol,
              and 2-Chlorophenol-d4 at a concentration of 150 ug/mL.   Store
              the spiking solutions at 4°C (±2°C) in Teflon-sealed
              containers.  The solutions should be checked frequently for
              stability.  These solutions must be replaced after twelve
              months, or sooner if comparison with quality control check
              samples indicates a problem.

      1.5.4   Base/Neutral and Acid Matrix Spiking solution

              Prepare a spiking solution in methanol that contains the
              following compounds at a concentration of 100 ug/mL for
              base/neutrals and 150 ug/mL for acids.  Store the spiking
              solutions at 4°C (±2°C) in Teflon-sealed containers.  The
              solutions should be checked frequently for stability.  These
              solutions must be replaced after twelve months, or sooner if
              comparison with quality control check samples indicates a
              problem.
                                 D-14/SV                      OLM01.1   12/90

-------
                                                              SECTION II
                     Bases/ Neutrals                    Acids

                     1,2,4-Trichlorobenzene           Pentachlorophenol
                     Acenaphthene                     Phenol
                     2,4-Dinitrotoluene               2-Chlorophenol
                     Pyrene                       .    4-Chloro-3-methylphenol
                     N-Nitroso-di-n-propylamine       4-Nitrophenol
                     1,4-Dichlorobenzene
1.6   Equipment
      1.6.1.  Glass  scintillation vials  -  at least 20 mL, with screw cap and
              teflon or  aluminum foil liner.

      1.6.2   Spatula -  stainless steel  or Teflon.

      1.6.3   Balance -  capable of weighing 100 g to + 0.01 g.

      1.6.4   Vials  and  caps  - 2 mL for  GC auto sampler.

      1.6.5   Disposable pipets - Pasteur; glass wool rinsed with methylene
              chloride.

      1.6.6   Concentrator tubes - 15 mL.

      1.6.7   Ultrasonic cell disrupter  - Heat Systems, Ultrasonics, Inc.,
              Model  W-385 SONICATOR (475 Watt with pulsing capability, No..
              200, 1/2 inch tapped disrupter horn and No. 419, 1/8  inch
              standard tapered MICROTIP  probe), or equivalent device with a
              minimum of 375 Watt output capability.  NOTE:  In order to
              ensure that sufficient energy is transferred to the sample
              during extraction, the MICROTIP probe must be replaced if the
              tip begins to erode.  Erosion of the tip is evidenced by a
              rough  surface.

      1.6.8   Sonabox acoustic enclosure - recommended with above disrupters
              for decreasing cavitation  sound.

      1.6.9   Test tube  rack.

      1.6.10  Oven -  drying.

      1.6.11  Desiccator.

      1.6.12  Crucibles  - porcelain.

      1.6.13  Syringes - 0.5 mL volume.

1.7   Medium Level Sample Preparation.

      1.7.1.  Transfer the sample container into a fume hood.  Decant and
              discard any water layer on a sediment sample.  Mix samples
              thoroughly, especially composited samples.  Discard any foreign
              objects such as sticks, leaves, and rocks.  Transfer

                                 D-15/SV                      OLM01.2   1/91

-------
                                                        SECTION II
         approximately 1 g  (record weight to the nearest 0.1 g) of
         sample  to a 20-mL vial.  Wipe the mouth of the vial with a
         tissue  to remove any sample material.  Record the exact weight
         of sample taken.  Cap the vial before proceeding with the next
         sample  to avoid any cross-contamination.

         1.7.1.1   Transfer 50 g of soil/sediment to a 100 mL beaker.
                   Add 50 mL of water and stir for 1 hour.  Determine
                   pH of sample with glass electrode and pH meter while
                   stirring.  Report pH value on appropriate data
                   sheets.  If the pH of the soil is greater than 11 or
                   less than 5, contact the Technical Project Officer
                   cited in the contract for instructions on how to
                   handle the sample.  Document the instructions in the
                   SDG Narrative.  Discard this portion of sample.
                   NOTE:  If limited sample volume is received,  use 5 g
                   of soil and 5 mL of water for the pH determination.
                   Note this in the SDG Narrative.

1.7.2    Immediately after weighing the sample for extraction, weigh
         5-10 g  of the sediment into a tared crucible.  Determine the
         percent moisture by drying overnight at 105°C.  Allow to cool
         in a desiccator before weighing.  Concentrations of individual
         analytes will be reported relative to the dry weight of
         sediment.

         g of sample - g of dry sample     1rt-
         £	5—=	^—=	±	s-—   x 100  = % moisture
                g of sample

1.7.3    Add 2.0 g of anhydrous powdered sodium sulfate to the sample  in
         the 20  mL vial from paragraph 1.7.1 and mix well.

1.7.4    Surrogates are added to all samples, spikes, and blanks.  Add
         0.5 mL  of surrogate spiking solution to sample mixture.

1.7.5    Add 0.5 mL of matrix standard spiking solution to, each of two 1
         g portions from the sample chosen for spiking.  The frequency
         of MS/MSD analysis is given in Section IV, paragraph 8.6.

1.7.6    Immediately add 9.5 mL of methylene chloride to the sample and
         disrupt the sample with the 1/8 inch tapered MICROTIP
         ultrasonic probe for 2 minutes at output control  setting 5,  in
         continuous mode (if using a sonicator other than Models W-375
         or W-385, contact the Projects Officer for appropriate output
         settings).  Before extraction, make certain that  the sodium
         sulfate is free flowing and not a consolidated mass.  As
         required, break up large lumps with a clean spatula or, very
         carefully, with the tip of the unenergized probe.

         Add only 9.0 mL of methylene chloride to the matrix spike
         samples to achieve a final volume of 10 mL.
                           D-16/SV                              OLM01.0

-------
                                                              SECTION II
      1.7.7    Prepare  a  method blank with each group of medium soil/sediment
               samples  extracted.   For  semivolatile analyses, a method blank
               for  medium soil/sediment samples consists of 1 g of sodium
               sulfate  (see  paragraph 1.5.1), spiked with the surrogates and
               carried  through the  entire analytical procedure.  The frequency
               of method  blank analysis is given in Section IV, paragraph 8.7

      1.7.8    Loosely  pack  disposable  Pasteur pipets with 2-3 cm glass wool
               plugs.   Filter the extract through the glass wool and collect
               at least 8.0  mL in a concentrator tube.

      1.7.9    If the extract is to be  screened prior to GPC, concentrate 5.0
               mL of the  extract collected in paragraph 1.7.7 to 1.0 mL using
               the  nitrogen  evaporation technique described in paragraph
               3.6.2.   Transfer the concentrate to an autosampler vial for
               GC/FID or  GC/MS for  screening.  The quantitation limits for the
               screening  procedure  in Section III are approximately 10,000
               ug/Kg.

      1.7.10   If the extract is to be  cleaned up using GPC without screening,
               take at  least 8.0 mL of  the extract in paragraph 1.7.7 and
               proceed  to paragraph 3 of this section.  Following GPC, the 5.0
               mL of extract collected  must be concentrated to 0.5 mL by the
               nitrogen evaporation technique described in paragraph 3.6.2,
               and  screened  according to the procedures in Section III.  In
               this case,  the quantitation limits for the screening procedures
               in Section III are approximately 20,000 ug/Kg.

2.     Low Level Preparation for Screening and Analysis of Semivolatiles

2.1   Summary of Method

      A 30 gram portion of sediment is  mixed with anhydrous powdered sodium
      sulfate and extracted with 1:1  methylene chloride/acetone using an
      ultrasonic probe.  If the  optional low level screen is used,  a portion
      of this dilute extract is  concentrated fivefold and is screened by
      GC/FID or GC/MS.   If peaks are  present at greater than 10,000 ug/kg,
      discard the extract and prepare the sample  by the medium level method.
      If no peaks are present at greater than 10,000 ug/kg,  the entire
      extract is concentrated,  subjected to GPC clean up,  and analyzed by
      GC/MS for extractable organics.

2.2   Interferences

      Method interferences may be caused by contaminants in solvents,
      reagents, glassware, and other  sample-processing hardware that lead to
      discrete artifacts  and/or  elevated baselines in the total ion current
      profiles.  All of these materials routinely must be demonstrated to be
      free from interferences under the conditions of the analysis by running
      laboratory reagent  blanks.   Matrix interferences may be caused by
      contaminants  that are coextracted from the  sample.   The extent of
      matrix interferences will  vary  considerably from source to source.
                                 D-17/SV                        OLM01.2   1/91

-------
                                                              SECTION  II


2.3   Apparatus and Materials

      Brand names,  suppliers, and part numbers  are  for  illustrative  purposes
      only.  No endorsement is implied.   Equivalent performance  may  be
      achieved using apparatus and materials  other  than those  specified here,
      but demonstration of equivalent performance meeting the  requirements  of
      this SOW is the responsibility of the Contractor.

      2.3.1   Apparatus for determining percent moisture

              2.3.1.1    Oven - drying.

              2.3.1.2    Desiccator.

              2.3.1.3    Crucibles - porcelain.

      2.3.2   Disposable Pasteur glass pipets  - 1 mL.

      2.3.3   Ultrasonic cell disrupter, Heat Systems,  Ultrasonics,  Inc.,
              Model W-385 SONICATOR  (475 Watt with pulsing capability, No.
              305, 3/4 inch  tapped high gain "Q" disrupter horn, or No. 208,
              3/4  inch standard solid disrupter horn),  or equivalent device
              with a minimum of 375 Watt output capability.  NOTE:  In order
              to ensure that sufficient energy is transferred to the sample
              during extraction, the horn must be replaced if the tip begins
              to erode.  Erosion of the tip is evidenced by a rough surface.

              Sonabox acoustic enclosure - recommended with above disruptors
              for decreasing cavitation sound.

      2.3.4   Beakers - 400 mL.

      2.3.5   Vacuum filtration apparatus.

              2.3.5.1    Buchner funnel.

              2.3.5.2    Filter paper - Whatman No. 41  or equivalent.

      2.3.6   Kuderna-Danish (K-D) apparatus.

              2.3.6.1    Concentrator tube -  10 mL, graduated (Kontes
                         K-570040-1025 or equivalent).

              2.3.6.2    Evaporative flask -  500 mL (Kontes K-570001-0500 or
                         equivalent).

              2.3.6.3    Snyder column - three-ball macro (Kontes
                         K-503000-0121 or equivalent).

              2.3.6.4    Snyder column - two-ball micro (Kontes K-569001-0219
                         or equivalent).
                                 D-18/SV   '                    OLM01.2   1/91

-------
                                                              SECTION II


      2.3.7   Silicon carbide boiling chips - approximately 10/40 mesh. Heat
              to 400°C for 30 minutes or Soxhlet extract with methylene
              chloride.

      2.3.8   Water bath - heated, with concentric ring cover, capable of
              temperature control (+2°C).  The bath should be used in a hood.

      2.3.9   Balance - capable of accurately weighing + 0.01 g.

      2.3.10  Vials and caps - 2 mL for GC auto sampler.

      2.3.11  Balance - analytical, capable of accurately weighing + O.OOOlg.

      2.3.12  Nitrogen evaporation device equipped with a water bath that can
              be maintained at 35-40"C.  (N-Evap by Organomation Associates,
              Inc., South Berlin, MA, or equivalent).

      2.3.13  Pyrex glass wool.

      2.3.14  Pasteur pipets - disposable.

      2.3.15  Syringes - 0.5 mL volume.

2.4   Reagents

      2.4.1   Sodium Sulfate - anhydrous powdered reagent grade, heated at
              400°C for four hours, cooled in a desiccator, and stored in a
              glass bottle (Baker anhydrous powder, catalog #73898 or
              equivalent).

      2.4.2   Methylene chloride, methanol, acetone, isooctane, 2-propanol,
              and benzene - pesticide residue analysis grade or equivalent.

      2.4.3   Reagent water - defined as water in which an interferent is not
              observed at or above the CRQL of each parameter of interest.

      2.4.4   Sodium Sulfit'e - reagent grade.

      2.4.5   Base/Neutral and Acid Surrogate Spiking Solution

              Surrogate standards are added to all samples and calibration
              solutions.  The compounds specified are Phenol-d5,
              2,4,6-Tribromophenol, 2-Fluorophenol, Nitrobenzene-d5,
              Terphenyl-di4, 2-Fluorobiphenyl, 2-Chlorophenol-d4, and 1,2-
              Dichlorobenzene-d4.  Prepare a surrogate standard spiking
              solution that contains Nitrobenzene-d5, Terphenyl-d^,
              2-Fluorobiphenyl, and l,2-Dichlorobenzene-d4 at a concentration
              of 100 ug/mL; Phenol-ds, 2,4,6-Tribromophenol, 2-Fluorophenol,
              and 2-Chlorophenol-d4 at a concentration of 150 ug/mL.   Store
              the spiking solutions at 4°C (±2°C) in Teflon-sealed
              containers.  The solutions should be checked frequently for
              stability.  These solutions must be replaced after twelve
              months, or sooner if comparison with quality control check
              samples indicates a problem.

                                 D-19/SV                       OLM01.2  1/91

-------
                                                              SECTION  II
      2.4.6   Base/Neutral and Acid Matrix Spiking solution

              Prepare a spiking solution in raethanol that contains the
              following compounds at a concentration of 100 ug/mL for
              base/neutrals and 150 ug/mL for acids.  Store the spiking
              solutions at 4°C (±2°C) in Teflon-sealed containers.  The
              solutions should be checked frequently for stability.  These
              solutions must be replaced after twelve months, or sooner if
              comparison with quality control check samples indicates a
              problem.

                     Bases/ Neutrals                    Acids

                     1,2,4-Trichlorobenzene            Pentachlorophenol
                     Acenaphthene                     Phenol
                     2,4-Dinitrotoluene                2-Chlorophenol
                     Pyrene                           4-Chloro-3-methylphenol
                     N-Nitroso-di-n-propylamine        4-Nitrophenol
                     1,4-Dichlorobenzene

2.5   Low Level  Sample Preparation

      2.5.1   Decant and discard any water layer on a sediment sample.  Mix
              samples thoroughly, especially composited samples.  Discard any
              foreign objects such as sticks, leaves, and rocks.

              Transfer 50 g of soil/sediment to 100 mL beaker.  Add 50 mL of
              water and stir for 1 hour.   Determine pH of sample with glass
              electrode and pH meter while stirring.  Report pH value on
              appropriate data sheets.  If the pH of the soil is greater than
              11 or less than 5, contact the Technical Project Officer cited
              in the contract for instructions on how to handle the sample.
              Document the instructions in the SDG Narrative.  Discard this
              portion of sample.  NOTE:   If limited sample volume is
              received, use 5 g of soil and 5 mL of water for the pH
              determination.  Note this in the SDG Narrative.

      2.5.2   The following steps should be performed rapidly to avoid loss
              of the more volatile extractables.  Weigh approximately  30 g of
              sample to the nearest 0.1 g into a 400-mL beaker and add 60 g
              of anhydrous powdered sodium sulfate.  Mix well.  The sample
              should have a sandy texture at this point.  Immediately, add
              100 mL of 1:1 methylene chloride-acetone to the sample,  then
              add the surrogates according to paragraph 2.5.2.3.

              2.5.2.1    Immediately after weighing the sample for
                         extraction, weigh 5-10 g of the sediment into a
                         tared crucible.   Determine the percent moisture by
                         drying overnight at 105°C.  Allow to cool in a
                         desiccator before weighing.  Concentrations of
                         individual analytes will be reported relative to the
                         dry weight of sediment.
                                 D-20/SV                              OLM01.0

-------
                   g of sample - g of dry sample
                   	•=	=	   x 100  = % moisture
                          g of sample

         2.5.2.2    Weigh out two 30 g (record weight to nearest 0.1 g)
                   portions for use as matrix and matrix spike
                   duplicates according to paragraph 2.5.2.  Add 0.5 mL
                   of the BNA matrix spike solution to each of two
                   portions.  The frequency of MS/MSD analysis is given
                   in Section IV, paragraph 8.6.

         2.5.2.3    Add 0.5 mL of base/neutral and acid surrogate
                   standard to the sample and each of the aliquots in
                   2.5.2.2.

         2.5.2.4    Prepare a method blank with each group of low
                   soil/sediment samples extracted.  For semivolatile
                   analyses, a method blank for low soil/sediment
                   samples consists of 30 g of sodium sulfate (see
                   paragraph 2.4.1), spiked with the surrogates and
                   carried through the entire analytical procedure.
                   The frequency of method blank analysis is given in
                   Section IV, paragraph 8.7

2.5.3    Place  the bottom surface of the tip of the 3/4 inch disrupter
         horn about 1/2 inch below the surface of the solvent but above
         the sediment layer.

2.5.4    Sonicate for 1 1/2 minutes with the W-385  (or 3 minutes with
         the W-375), using No. 208, 3/4 inch standard disrupter horn
         with output control knob set at 10 (or No. 305, 3/4 inch tapped
         high gain "Q" disruptor horn at 5) and mode switch on "1 sec.
         pulse" and % duty cycle knob set at 50% (if using a sonicator
         other  than Models W-375 or W-385, contact  the Project Officer
         for appropriate output settings).  Do NOT  use MICROTIP probe.

2.5.5    Decant and filter extracts through Whatman #41 filter paper
         using vacuum filtration or centrifuge and  decant extraction
         solvent.

2.5.6    Repeat the extraction two more times with  2 additional 100 mL
         portions of 1:1 methylene chloride-acetone.  Before each
         extraction, make certain that the sodium sulfate is free
         flowing and not a consolidated mass.  As required, break up
         large  lumps with a clean spatula or, very  carefully, with the
         tip of the unenergized probe.  Decant the  extraction solvent
         after  each sonication.  On the final sonication, pour the
         entire sample into the Buchner funnel and  rinse with 1:1
         methylene chloride-acetone.
                           D-21/SV                        OLM01.2   1/91

-------
               2.5.6.1    If the sample is to be screened from the low level
                         method prior to GPC, take 5.0 mL and concentrate to
                         1.0 mL following paragraph 3.6.1 or 3.6.2,  but note
                         that the final volume for screening is 1.0  mL, not
                         0.5 mL.  Screen the extract as per Section  III,
                         paragraph 1., "Screening of Extractable Organic
                         Extracts."

               2.5.6.2    After GC/FID or GC/MS screening, transfer the
                         remainder of the 1 mL back to the total extract from
                         paragraph 2.5.6.  CAUTION: To minimize sample loss,
                         autosamplers which pre-flush samples through the
                         syringe should not be used.

2.6   Concentration and Solvent Exchange

      2.6.1    Low level soil/sediment samples prepared by the procedures in
               paragraph 2.5 will result  in extracts containing a mixture of
               acetone and methylene chloride.  Because all soil/sediment
               sample extracts must be subjected to GPC clean up prior to
               analysis, the majority of  the acetone must be removed from the
               extract, otherwise it will have adverse effects on the GPC
               column.  To remove the acetone from the sample extract, follow
               the steps in 2.6.2-2.6.4.

      2.6.2    Transfer the extract to a Kuderna-Danish (K-D) concentrator
               consisting of a 10 mL concentrator tube and a 500 mL
               evaporative flask.  Other concentration devices or techniques
               may be used if equivalency is demonstrated for all extractable
               compounds listed in Exhibit C.

      2.6.3   Add one or two clean boiling chips to the evaporative  flask and
               attach a three-ball Snyder column.  Pre-wet the Snyder column
              by adding about 1 mL methylene chloride to the top. Place the
              K-D apparatus on a hot water bath (60 to 80°C) so that the
              concentrator tube is partially immersed in the hot water and
               the entire lower rounded surface of the flask is bathed with
              hot vapor.  Adjust the vertical position of the apparatus and
               the water temperature as required to complete the concentration
               in 10 to 15 minutes.  At the proper rate of distillation the
              balls of the column will actively chatter but the chambers will
              not flood with condensed solvent.  When the apparent volume of
               liquid reaches 1 mL, remove the K-D apparatus and allow it to
               drain and cool for at least 10 minutes.  Do not allow the
              evaporator to go dry.

      2.6.4   Dilute the extract to 10.0 mL with methylene chloride, and
              proceed with GPC clean up  (see paragraph 3).
                                 D-22/SV                       OLM01.2  1/91

-------
3.    Extract Cleanup by Gel Permeation Chromatography

      Gel permeation chromatography (GPC) is a size exclusion cleanup
      procedure using organic solvents and hydrophobic gels in the separation
      of synthetic macromolecules.   The packing gel is porous and is
      characterized by the range or uniformity (exclusion range) of that pore
      size.  In the choice of gels, the exclusion range must be larger than
      the molecular size of the molecules to be separated.  A cross-linked
      divinyl benzenestyrene copolymer (SX-3 Bio Beads or equivalent) is
      specified for this method.

      GPC is required for all soil/sediment samples,  regardless of
      concentration level, for the elimination of lipids, polymers,
      copolymers, proteins, natural resins and polymers, cellular components,
      viruses, steroids, and dispersed high-molecular-weight compounds from
      the sample extract.  GPC is appropriate for both polar and non-polar
      analytes, therefore, it can be used effectively to clean up extracts
      containing a broad range of analytes.

      Normally, this method is most efficient for removing high boiling
      materials that condense in the injection port area of a gas
      chromatograph (GC) or in the front of the GC column.  This residue
      ultimately will reduce the chromatographic separation efficiency or
      column capacity because of adsorption of the target analytes on the
      active sites.  Pentachlorophenol especially is susceptible to this
      problem.

      In the event that the Laboratory fails to appropriately employ GPC
      clean-up procedures, the Agency will require the clean up and
      reanalysis of all affected samples or sample extracts at no additional
      cost to the Agency.

3.1   Apparatus and Materials

      Brand names, suppliers, and part numbers are for illustrative purposes
      only.   No endorsement is implied.  Equivalent performance may be
      achieved using apparatus and materials other than those specified here,
      but demonstration of equivalent performance meeting the requirements of
      this SOW is the responsibility of the Contractor.

      3.1.1   Gel permeation chromatography  (GPC) cleanup device.  NOTE: GPC
              cleanup is  required  for  all soil/sediment  extracts.

              Gel permeation chromatography  system - GPC Autoprep Model  1002
              A or B (Analytical Biochemical Laboratories, Inc., or
              equivalent) Systems  that perform very satisfactorily also  have
              been assembled from  the  following components - an HPLC  pump,  an
              auto sampler or a valving system with sample loops, and a
              fraction collector.  All systems, whether  automated or  manual,
              must meet the calibration requirements of  paragraph 3.4.
                                 D-23/SV                       OLM01.2   1/91

-------
                                                              SECTION II
               3.1.1.1    Chromatographic column - 700 mm x 25 mm i.d. glass
                         column.  Flow is'upward.  To simplify switching from
                         the UV detector during calibration to the GPC
                         collection device during extract cleanup, an
                         optional double 3-way valve (Rheodyne Type 50 Teflon
                         Rotary Valve #10-262 or equivalent) may be attached
                         so that the column exit flow can be shunted either
                         to the UV flow-through cell or to the GPC collection
                         device.

               3.1.1.2    Guard column - (Optional) 5 cm, with appropriate
                         fittings to connect to the inlet side of the
                         analytical column (Supelco 5-8319 or equivalent).

               3.1.1.3    Bio Beads (S-X3) - 200-400 mesh, 70 gm (Bio-Rad
                         Laboratories, Richmond, CA, Catalog 152-2750 or
                         equivalent).  An additional 5 gm of Bio Beads is
                         required if the optional guard column is employed.
                         The quality of Bio Beads may vary from lot to lot
                         because of excessive fines in some lots.  In
                         addition to fines having a detrimental effect on
                         chromatography, they also can pass through the
                         column screens and damage the valve.

               3.1.1.4    Ultraviolet detector - fixed wavelength (254 nm)
                         with a semi-prep flow-through cell.

               3.1.1.5    Strip chart recorder, recording integrator or
                         laboratory data system.

               3.1.1.6    Syringe - 10-mL with Luerlok fitting.

               3.1.1.7    Syringe filter assembly, disposable - Bio-Rad "Prep
                         Disc" sample filter assembly #343-0005, 25 mm, and 5
                         micron filter discs or equivalent.  Check each batch
                         for contaminants.   Rinse each filter assembly (prior
                         to use) with methylene chloride if necessary.

               3.1.1.8    A description of a manual system assembled from
                         parts can be found in Wise, R.H., Bishop, D.F.,
                         Williams., R.T. & Austern, B.M. "Gel Permeation
                         Chromatography in the GC/MS Analysis of Organics  in
                         Sludges" U.S. EPA, Municipal Environmental Research
                         Laboratory, Cincinnati, Ohio, 45268.
3.2   Reagents
      3.2.1   GPC Calibration Solution  - prepare  a calibration solution in
              methylene chloride containing  the following  analytes  (in
              elution order):
                                  D-24/SV                              OLM01.0

-------
                                                              SECTION II
                   Compound                            mg/mL
                   corn oil                            25.0
                   bis(2-ethylhexyl)phthalate           1.0
                   me thoxychlor                         0.2
                   perylene                             0.02
                   sulfur (optional)              .      0.08

              NOTE:  If used, sulfur is not very soluble in methylene
              chloride, however, it is soluble in warm corn oil.  Therefore,
              one approach is to weigh out the corn oil, warm it and transfer
              the weighed amount of sulfur into the warm corn oil.  Mix it
              and then transfer into a volumetric flask with methylene
              chloride, along with the other calibration compounds.

              Store the calibration solution in an amber glass bottle with a
              Teflon-lined screw-cap at 4°C, and protect from light
              (refrigeration may cause the corn oil to precipitate.  Before
              use, allow the calibration solution to stand at room
              temperature until the corn oil dissolves).  Replace the
              calibration standard solution every 6 months, or more
              frequently if necessary.

3.3   Column Preparation

      3.3.1   Weigh out 70 gm of Bio Beads (SX-3).  Transfer them to a quart
              bottle with a Teflon-lined cap or a 500 mL separatory funnel
              with a large bore stopcock, and add approximately 300 mL of
              methylene chloride.  Swirl the container to ensure the wetting
              of all beads.  Allow the beads to swell for a minimum of 2
              hours.  Maintain enough solvent to cover the beads sufficiently
              at all times.  If a guard column is to be used, repeat the
              above with 5 gm of Bio Beads in a 125 mL bottle or a beaker,
              using 25 mL of methylene chloride.

      3.3.2   Turn the column upside down from Its normal position, and
              remove the inlet bed support plunger (the inlet plunger is
              longer than the outlet plunger).  Position and tighten the
              outlet bed support plunger as near the end as possible, but no
             . closer than 5 cm (measured from the gel packing to the collar).

      3.3.3   Raise the end of the outlet tube to keep the solvent in the GPC
              column, or close the column outlet stopcock.  Place a small
              amount of solvent in the column to minimize the formation of
              air bubbles at the base of poured column packing.

      3.3.4   Swirl the bead/solvent slurry to get a homogeneous mixture and,
              if the wetting was done in a quart bottle, quickly transfer it
              to a 500 mL separatory funnel with a large bore stopcock.
              Drain the excess methylene chloride directly into the waste
              beaker, and then start draining the slurry into the column by
              placing the separatory funnel tip against the column wall.
              This will help to minimize bubble formation.  Swirl
              occasionally Tzo keep the slurry homogeneous.  Drain enough to
              fill the column.  Place the tubing from the column outlet into

                                 D-25/SV                             OLM01.0

-------
                                                        SECTION II
        .a waste beaker below the column, open the stopcock (if
        attached), and allow the excess solvent to drain.  Raise the
        tube  to stop the flow, and close the stopcock when the top of
        the gel begins to look dry.  Add additional methylene chloride
        to just rewet the gel.

3.3.5   Wipe  any remaining beads and solvent from the inner walls of
        the top of the column with a laboratory tissue.  Loosen the
        seal  slightly on the other plunger assembly (long plunger) and
        insert it into the column.  Make the seal just tight enough so
        that  any beads on the glass surface will be pushed forward, but
        loose enough so that the plunger can be pushed forward.

        CAUTION:  Do not tighten the seal if beads are between the seal
        and the glass surface because this can damage the seal and
        c aus e 1eakage.

3.3.6   Compress the column as much as possible without applying
        excessive force.  Loosen the seal and gradually pull out the
        plunger.  Rinse and wipe off the plunger.  Slurry any remaining
        beads and transfer them into the column.  Repeat the step in
        paragraph 3.3.5 and reinsert the plunger.  If the plunger
        cannot be inserted and pushed in without allowing beads to
        escape around the seal, continue compression of the beads
        without tightening the seal, and loosen and remove the plunger
        as described.  Repeat this procedure until the plunger is
        inserted successfully.

3.3.7   Push  the plunger until it meets the gel, then compress the
        column bed about four centimeters.

3.3.8   Pack  the optional 5 cm column with approximately 5 gm of
        preswelled beads (different guard columns may require different
        amounts).  Connect the guard column to the inlet of the
        analytical column.

3.3.9   Connect the column inlet to the solvent reservoir (reservoir
        should be placed higher than the top of the column) and place
        the column outlet tube in a waste container.  Placing a
        restrictor in the outlet tube will force air out of the column
        more  quickly.  A restrictor can be made from a piece of
        capillary stainless steel tubing of 1/16" OD x 10/1000" ID x
        2".   Pump methylene chloride through the column at a rate of 5
        mL/min for one hour.

3.3.10  After washing the column for at least one hour, connect the
        column outlet tube, without the restrictor, to the inlet side
        of the UV detector.  Connect the system outlet to the outlet
        side  of the UV detector.  A restrictor (same size as the one in
        paragraph 3.3.9) in the outlet tube from the UV detector will
        prevent bubble formation which causes a noisy UV baseline.  The
        restrictor will not effect flow rate.  After pumping methylene
                           D-26/SV                              OLM01.0

-------
                                                              SECTION  II
              chloride through the column for an additional 1-2 hours, adjust
              the inlet bed support plunger until approximately 6-10 psi
              backpressure is achieved.  Push the plunger in to increase
              pressure or slowly pull outward to reduce pressure.

      3.3.11  When the GPC column is not to be used for several days, connect
              the column outlet line to the column inlet to prevent column
              drying and/or channeling.  If channeling occurs, the gel must
              be removed from the column, reswelled, and repoured as
              described above.  If drying occurs, methylene chloride should
              be pumped through the column until the observed column pressure
              is constant and the column appears wet.  Always recalibrate
              after column drying has occurred to verify retention volumes
              have not changed.

3.4   Calibration of the GPC Column

      3.4.1   Using a 10 mL syringe, load sample loop #1 with calibration
              solution (paragraph 3.2).  With the ABC automated system, the 5
              mL sample loop requires a minimum of 8 mL of the calibration
              solution.  Use a firm, continuous pressure to push the sample
              onto the loop.  Switch the valve so that GPC flow is through
              the UV flow-through cell.

      3.4.2   Inject the calibration solution and obtain a UV trace showing a
              discrete peak for each component.  Adjust the detector and/or
              recorder sensitivity to produce a UV trace that meets the
              following requirements.  Differences between manufacturer's
              cell volumes and detector sensitivities may require a dilution
              of the calibration solution to achieve similar results.  An
              analytical flow-through detector cell will require a much less
              concentrated solution than the semi-prep cell and, therefore,
              the analytical cell is not acceptable for use.

              o   Peaks must be observed and should be  symmetrical  for all
                  compounds in the calibration solution.
              o   Corn oil  and phthalate peaks must exhibit >85% resolution.

              o   Phthalate and methoxychlor peaks  must exhibit >85%
                  resolution.

              o   Methoxychlor and perylene  peaks must  exhibit >85%
                  resolution.

              o   Perylene  and sulfur peaks  must not be saturated and must
                  exhibit >90% baseline resolution.

      3.4.3   Using the information from the UV trace,  establish appropriate
              collect and dump time periods to ensure collection of all
              target analytes.  Initiate column eluate collection just before
              elution of bis(2-ethylhexyl)phthalate and after the elution of
                                 D-27/SV                             OLM01.0

-------
                                                        SECTION II
         the  corn oil.   Stop eluate collection shortly after the elution
         of perylene.   Collection should be stopped before sulfur
         elutes.   Use a "wash"  time of  10 minutes after the elution of
         sulfur.   Each  laboratory is required to establish its specific
         time sequences.

3.4.4    Verify  the  flow rate by collecting column eluate for 10 minutes
         in a graduated cylinder and measure the volume, which should be
         45-55 mL (4.5-5.5 mL/min).  If the flow rate is outside of this
         range,  corrective action must  be taken, as described above.
         Once the flow  rate is  within the range of 4.5-5.5 mL/min,
         record  the  column pressure (should be 6-10 psi) and room
         temperature.   Changes  in pressure, solvent flow rate, and
         temperature conditions can affect analyte retention times and
         must be  monitored.  If the flow rate and/or column pressure do
         not  fall within the above ranges, a new column should be
         prepared.   A UV trace  that does not meet the criteria in
         paragraph 3.4.2 would  also indicate that a new column should be
         prepared.   It  may be necessary to obtain a new lot of Bio Beads
         if the  column  fails all the criteria.

3.4.5    Reinject the calibration solution after appropriate collect and
         dump cycles have been  set, and the solvent flow and column
         pressure have  been established.

         3.4.5.1     Measure and record  the volume of collected GPC
                    eluate in a graduated cylinder.  The volume of GPC
                    eluate collected for each sample extract processed
                    may be used to indicate problems with the system
                    during sample processing.

         3.4.5.2     The  retention times for bis(2-ethylhexyl)phthalate
                    and perylene must not vary more than +5% between
                    calibrations.  If the retention time shift is >5%,
                    take corrective action.   Excessive retention time
                    shifts are  caused by the following:

                    o    Poor laboratory temperature control or system
                        leaks.

                    o    An unstabilized column that requires pumping
                        methylene chloride through it for several more
                        hours or overnight.

                    o    Excessive laboratory temperatures causing
                        outgassing of the methylene chloride.

3.4.6    Analyze  a GPC  blank by loading 5 mL of methylene chloride into
         the  GPC.  Concentrate  the methylene chloride that passes
         through  the system during the  collect cycle using a Kuderna-
         Danish  (KD) evaporator.  Analyze the concentrate by GC/MS.  If
         the  blank exceeds one  half the CRQL of any analyte, pump
         additional  methylene chloride  through the system for 1-2 hours.
         Analyze  another GPC blank to ensure the system is sufficiently
         clean.   Repeat the methylene chloride pumping if necessary.

                           D-28/SV                              OLM01.0

-------
                                                              SECTION  II
3 . 5   Sample Extract Cleanup

      It is very important to have consistent  laboratory temperatures  during
      an entire GPC run,  which could be 24  hours  or  more.   If  temperatures
      are not consistent,  retention times will shift,  and the  dump  and
      collect times determined by the calibration standard no  longer will be
      appropriate.   The ideal laboratory temperature to  prevent  outgassing  of
      the methylene chloride is 22'C.

      3.5.1   In order to prevent overloading of the GPC column, highly
              viscous sample extracts must be diluted prior to cleanup.  Any
              sample extract with a viscosity greater than that of a 1:1
              glycerol:water solution must be diluted and loaded into several
              loops.  Similarly, extracts containing more than 500 mg of
              nonvolatile residue per 5 mL of extract must be diluted and
              loaded into several loops.  The nonvolatile residue may be
              determined by evaporating a 100 uL aliquot of the extract to
              dryness in a tared aluminum weighing pan,  or other suitable
              container.

      3.5.2   Particles greater than 5 microns may scratch the valve,  which
              may result in a system leak and cross contamination of sample
              extracts in the sample loops.  To avoid such problems,  filter
              the extract through a 5 micron filter disc by attaching a
              syringe filter assembly containing the filter disc to a 10 mL
              syringe.  Draw the sample extract through  the filter assembly
              and into the 10 mL syringe.   Disconnect the filter assembly
              before transferring the sample extract into a small glass
              container, e.g., a 15 mL culture tube with a Teflon lined screw
              cap.   Alternatively, draw the extract into the syringe without
              the filter assembly.  Attach the filter assembly and force the
              extract through the filter and into the glass container.  Draw
              a minimum of 8 mL of extract into a 10 mL  syringe.

      3.5.3   Attach the syringe to the turn lock on the injection port.  Use
              firm, continuous pressure to push the sample onto the 5-mL
              sample loop.  If the sample is difficult'to load,  some part of
              the system may be blocked.  Take appropriate corrective action.
              If the back pressure is normal (6-10 psi)  the blockage is
              probably in the valve.  Blockage may be flushed out of the
              valve by reversing the inlet and outlet tubes and pumping
              solvent through the tubes (this should be  done before sample
              loading).

              NOTE:  Approximately 2 mL of the extract remains in the lines
              between the injection port and the sample  loop;  excess sample
              also  passes through the sample loop to waste.

      3.5.4   After loading a loop, and before removing  the syringe from the
              injection port, index the GPC to the next  loop.   This will
              prevent loss of sample caused by unequal pressure in the loops.
                                 D-29/SV                       OLM01.2  1/91

-------
                                                              SECTION II
      3; 5.5   -After loading each sample loop, wash the loading port with
              methylene chloride in a PTFE wash bottle to minimize cross
              contamination.  Inject approximately 10 mL of methylene
              chloride to rinse the common tubes.

      3.5.6   After loading all the sample loops,.index the GPC to the 00
              position, switch to the "RUN" mode and start the automated
              sequence.  Process each sample using the collect and dump cycle
              times established in 3.4.

      3.5.7   Collect each sample in a 250-mL Erlenmeyer flask, covered with
              aluminum foil to reduce solvent evaporation, or directly into a
              Kuderna-Danish evaporator.  Monitor sample volumes collected.
              .Changes in sample volumes collected may indicate one or more of
              the following problems:

              o   Change in solvent flow rate,  caused by channeling in the
                  column or changes in column pressure.
              o   Increase in column operating pressure  due to the absorption
                  of particles or gel  fines onto either  the guard column or
                  the analytical column gel,  if a  guard  column is not used.

              o   Leaks in the system  or significant variances in room
                  temperature.

      3.5.8   Concentrate the extract as per paragraphs 3.6.1 or 3.6.2.

      3.5.9   Calibrate the GPC at least once per week,  following the
              procedure outlined in 3.4.  The UV trace must meet the
              requirements in paragraph 3.4.2.  In addition,  the retention
              times of the calibration compounds must be within +5% of their
              retention times in the previous calibration.  A copy of the UV
              trace of the calibration solution must be submitted with the
              data for the associated samples.

      3.5.10  If the requirements in paragraphs 3.4.2 and 3.5.9 cannot be
              met, the column may be cleaned by processing several 5 mL
              volumes of butyl chloride throught the system.  Butyl chloride
              removes the discoloration and particles that may have
              precipitated out of the methylene chloride extracts.  If a
              guard column is being used, replace it with a new one.  This
              may correct the problem.  If column maintenance does not
              restore the performance of the column, the column must be
              repacked with new packing and recalibrated.

3.6   Final Concentration of Extract

      3.6.1   Transfer the sample extract to a K-D evaporator, attach the
              micro-Snyder column to the concentrator tube and add a silicon
              carbide boiling chip to the concentrator tube.  Pre-wet the
              Snyder column with 0.5 mL of methylene chloride.  Place the  K-D
              apparatus on the hot water bath  (80-90°C) so that the
              concentrator tube is partially immersed in the hot water.
              Adjust the vertical position of  the apparatus and the water

                                 D-30/SV                             OLM01.0

-------
                                                              SECTION II
               temperature as required to complete the concentration in 5 to
               10 minutes.  When the apparent volume of the liquid reaches 0.4
               mL, remove the K-D apparatus from the water bath and allow it
               to drain for at least 10 minutes while cooling.  Remove the
               Snyder column and rinse the lower joint into the concentrator
               tube with 0.1 mL of. methylene chloride.  Adjust the final
               volume to 0.5 mL with methylene chloride.  Concentrating the
               extract to 0.5 mL will result in no loss of sensitivity despite
               the volume of extract (5 mL) not recovered after GPC.

      3.6.2    Nitrogen evaporation technique (taken from ASTM Method D 3086)

               The following method may be used for final concentration of the
               semivolatile extract instead of the procedures in paragraph
               3.6.1.  Place the concentrator tube in a warm water bath (35°C)
               and evaporate the solvent volume to below 0.5 mL using a gentle
               stream of clean, dry nitrogen (filtered through a column of
               activated carbon).  CAUTION:  New plastic tubing must not be
               used between the carbon trap and the sample, since it may
               introduce interferences.

               The internal wall of the tube must be rinsed down several times
               with methylene chloride during the operation.  During
               evaporation, the tube solvent level must be kept below the
               water level of the bath.  The extract must never be allowed to
               become dry.  Concentrating the extract to 0.5 mL will result in
               no loss of sensitivity despite the volume of extract (5 mL) not
               recovered after GPC.

3.7   Store all extracts  at  4°C  (±2°C)  in the  dark in Teflon-sealed
      containers.

3.8   If the extract was  not screened prior to GPC,  proceed to Section III
      for the screening procedures.   If the extract  was  screened prior to
      GPC,  proceed with the  GC/MS  analysis in  Section IV.
                                 D-31/SV                       OLM01.2   1/91

-------
       SECTION III
SCREENING OF SEMIVOLATILE
    ORGANIC EXTRACTS
        D-32/SV                             OLM01.0

-------
                                                             SECTION III
1.     Summary of Method

1.1   The solvent extracts of water and soil/sediment are screened on a gas
      chromatograph/flame ionization detector (GC/FID) using a fused silica
      capillary column (FSCC).  For water samples,  the results of the screen
      may be used to determine an appropriate dilution factor for the GC/MS
      analysis of the sample extract.  For soil/sedimant samples, the results
      of the screen are used to determine which of  the two sample preparation
      procedures (low or medium) is required, and to determine an appropriate
      dilution factor for GC/MS analysis.  The results of the screen may be
      used also to assist the analyst in performing Gel Permeation
      Chromatography (GPC) clean up procedures on extracts of either water or
      soil/sediment samples.

2.     Apparatus and Materials

2.1   Gas chromatograph - an analytical system complete with a temperature
      programmable gas chromatograph and all required accessories including
      syringes, analytical columns, and gases.  The injection port must be
      designed for on-column injection when using packed columns and for
      splitless injection when using capillary columns.

      2.1.1   Above GC equipped with flame ionization  detector.

      2.1.2   GC column  - 30  m x 0.32 mm, 1 micron film thickness, silicone
              coated, fused silica capillary  column (J & W Scientific DB-5
              or equivalent).

3.     Reagents

3.1   Methylene chloride - pesticide residue analysis grade or equivalent.

3..2   GC calibration standard - prepare a standard solution containing
      phenol,  phenanthrene and di-n-octylphthalate.

      3.2.1   Stock standard  solutions  (1.00  ug/uL) -  Stock  standard
              solutions can be prepared  from  pure  standard materials or
              purchased solutions.

              3.2.1.1    Prepare stock standard solutions by accurately
                         weighing  about 0.0100 g of pure material.  Dissolve
                         the material in pesticide quality methylene
                         chloride  and dilute  to volume in a 10 mL volumetric
                         flask.  Larger volumes may be used at the
                         convenience of the analyst.   If compound purity  is
                         assayed at 97% or greater, the weight may be used
                         without correction to calculate the concentration
                         of the stock standard.  Commercially prepared stock
                         standards may be used at any  concentration if they
                         are certified by the manufacturer or by an
                         independent source  (see Exhibit E).
                                 D-33/SV                        OLM01.2   1/91

-------
                                                             SECTION III


               3.2.1.2    Transfer the stock standard solutions into Teflon-
                         sealed screw-cap bottles.  Store at -10°C to -20°C
                         and protect from light.  Stock standard solutions
                         should be checked frequently for signs of
                         degradation or evaporation, especially just prior
                         to preparing calibration.standards from them.
                         Stock standard solutions must be replaced after six
                         months, or sooner, if comparison with quality
                         control check samples indicates a problem.
                         Standards prepared from gases or reactive compounds
                         such as styrene must be replaced after two months,
                         or sooner, if comparison with quality control check
                         samples indicates a problem.

      3.2.2    Prepare  a working standard mixture of the three compounds in
               methylene chloride.  The concentration must be such that  the
               volume injected equals 50 ng of each compound.  The storage
               and stability requirements are the same as specified in
               paragraph 3.2.1.2.

4.    GC Calibration

4.1   At the beginning of each 12 hour shift,  inject the GC calibration
      standard.  The following criteria must be  met:

      4.1.1    The GC must be standardized for half scale response from  50 ng
               of phenanthrene.

      4.1.2    The GC must adequately separate phenol from the solvent front.

      4.1.3    A  minimum of quarter scale response for 50 ng of
               di-n-octylphthalate must be exhibited.

5.    GC/FID Screening

5.1   Suggested GC operating conditions are as follows:

      o   Initial Column Temperature Hold - 50°C for 4 minutes.

      o   Column Temperature Program - 50-280°C  at 8 degrees/rain.

      o   Final Column Temperature Hold -  280°C  for 8 minutes.

      o   Injector - Grob-type,  splitless.

      o   Sample Volume -  1-2 uL.

      o   Carrier Gas - Helium at 30 mL/sec.

5.2   Inject the GC calibration standard and ensure the  criteria specified in
      4 are met before injecting samples.   Estimate the  response for 10 ng of
      phenanthrene.

5.3   Inject the appropriate extracts from Section II,  including blanks.

                                 D-34/SV                       OLM01.2  1/91

-------
                                                             SECTION III
6.    Interpretation of Chromatograms

6.1   Water

      6.1.1    If no  sample peaks are detected, or all are less than full
               scale  deflection, the undiluted extract is analyzed on GC/MS.

      6.1.2    If any sample peaks are greater than full scale deflection,
               calculate  the dilution necessary to reduce the major peaks  to
               between half and  full scale deflection.  Use this dilution
               factor to  dilute  the extract for GC/MS analysis.

6.2   Soil/Sediment

      6.2.1    If no  sample peaks from the extract (from low or medium level
               preparation) are  detected, or all are less than 10% full scale
               deflection, the sample must be prepared by the low level
               protocol,  Section II, Part C, beginning at 2.

      6.2.2    Peaks  are  detected at greater than 10% full scale deflection
               and  less than or  equal to full scale deflection.

               6.2.2.1   If the screen is from the medium level extract,
                         proceed with GC/MS analysis of this extract with
                         appropriate dilution if necessary.

               6.2.2.2   If screen is from the low level extract, discard
                         extract and prepare sample by medium level method
                         for GC/MS analysis.

      6.2.3    Peaks  are  detected at greater than full scale deflection.

               6.2.3.1   If the screen is from the medium level preparation,
                         calculate the dilution necessary to reduce the
                         major peaks to between half and full scale
                         deflection.  Use this dilution factor to dilute  the
                         extract.  This dilution is analyzed by GC/MS for
                         extractable organics.

               6.2.3.2   If the screen is from the low level preparation,
                         discard the extract and prepare a sample by the
                         medium level method for GC/MS analysis.

7.    GC/MS Analysis

7.1   Use the information from paragraph 6 (Interpretation of Chromatograms)
      to perform the GC/MS analysis, beginning Section IV,  GC/MS Analysis of
      Semivolatiles.

7.2   The information from paragraph 6 may be usefull also in processing
      sample extracts through GPC cleanup.

      NOTE:  The choice of screening sample extracts before or after GPC
      cleanup is left to the laboratory.

                                 D-35/SV                        OLM01.3   2/91

-------
           SECTION  IV
GC/MS ANALYSIS OF SEMIVOLATILES
          D-36/SV                              OLM01.0

-------
                                                              SECTION IV
1.     Summary of Method

      This method is to be used for the GC/MS  analysis  of semivolatiles
      screened by Section III protocols and for confirmation of
      pesticides/Aroclors identified by GC/EC,  if concentrations  permit.

2.     Apparatus and Materials

      Brand names,  suppliers, and part numbers are for  illustrative purposes
      only.  No endorsement is implied.  Equivalent performance may be
      achieved using apparatus and materials other than those specified  here,
      but demonstration of equivalent performance meeting the requirements  of
      this SOW is the responsibility of the Contractor.

2.1   Gas chromatograph/mass spectrometer system.

      2.1.1   Gas chromatograph  - an analytical system complete with a
              temperature programmable gas chromatograph suitable for
              splitless  injection and  all required accessories including
              syringes,  analytical columns and gases.

      2.1.2   Column  - 30 m  x 0.25 mm  ID (or 0.32 mm)  bonded-phase silicone
              coated  fused silica capillary column (J&W Scientific DB-5 or
              equivalent).   A film thickness of 1.0 micron is recommended
              because of its larger capacity.  A film thickness of 0.25
              micron may be  used.

      2.1.3   Mass Spectrometer  - capable of scanning from 35 to 500 amu
              every 1 second or  less,  utilizing 70 volts (nominal) electron
              energy  in  the  electron impact ionization mode and producing a
              mass spectrum  which meets all the instrument performance
              criteria in Table  1 when 50 ng of decafluorotriphenylphosphine
              (DFTPP) is injected through the GC inlet.  The instrument
              conditions required for  the acquisition of the DFTPP mass
              spectrum are given in  paragraph 4.3.4.   NOTE:  DFTPP criteria
              must be met before any sample extracts are analyzed.  Any
              samples analyzed when DFTPP criteria have not been met will
              require reanalysis at no cost to the Agency.

      2.1.4   GC/MS interface -  any gas chromatograph to mass spectrometer
              interface  that gives acceptable calibration points, at 50 ng or
              less per injection, for  each of the parameters of interest, and
              achieves all acceptable  performance criteria  (Exhibit E), may
              be used.  Gas  chromatograph to mass spectrometer interfaces
              constructed of all-glass or glass-lined materials are
              recommended.   Glass can  be deactivated by silanizing with
              dichlorodimethyls ilane.

      2.1.5   Data system -  a computer system must be interfaced to the mass
              spectrometer that  allows the continuous acquisition and storage
              on machine readable media of all mass spectra obtained
              throughout the duration  of the chromatographic program.  The
              computer must  have software that allows searching any GC/MS
              data file  for  ions of a  specific mass and plotting such ion

                                 D-37/SV                       OLM01.2   1/91

-------
                                                              SECTION IV
               abundances versus time or scan number.  This type of plot is
               defined  as an Extracted Ion Current Profile (EICP).  Software
               must be  available that allows integrating the abundance in any
               EICP between specified time or scan number limits'.  Also, for
               the non-target compounds, software must be available that
               allows comparing sample spectra against reference spectra.  The
               most recent release of the NIST/EPA/MSDC mass spectral library
               shall be used as the reference library.  The data system must
               be capable of flagging all data files that have been edited
               manually by laboratory personnel.

      2.1.6    Syringes - 2 uL and 10 uL volumes.

3.     Reagents

3.1   Internal standards -  1,4 Dichlorobenzene-d^,  Naphthalene-dg,
      Acenaphthene-diQ, Phenanthrene-d^g,  Chrysene-d^2•  Perylene-d]^ •

      An internal standard solution can be prepared by dissolving 100  mg of
      each compound in 50 mL of methylene  chloride.   It may  be necessary to
      use 5 to 10 percent benzene  or toluene in this solution and a few
      minutes of ultrasonic mixing in order to dissolve all  the constituents.
      The resulting solution will  contain  each standard at a concentration of
      2000 ng/uL.   A 10 uL portion of this solution should be added to each 1
      mL of sample extract.   This  will result in 40 ng of each internal
      standard in the 2 uL volume  of extract injected into the GC/MS.

3.2   Prepare calibration standards at a minimum of five concentration levels
      (20,  50, 80,  120, and 160 total ng per 2 uL).   Each calibration
      standard should contain each compound of interest and  each surrogate.
      Eight compounds,  2,4-Dinitrophenol,  2,4,5-Trichlorophenol,  2-
      Nitroaniline,  3-Nitroaniline,  4-Nitroaniline,  4-Nitrophenol,  4,6-
      Dinitro-2-methylphenol,  and  Pentachlorophenol will require only  a four-
      point initial calibration at 50,  80,  120,  and 160 total ng,  since
      detection at less that 50 ng per injection is difficult.  Great  care
      must be taken to maintain the integrity of all standard solutions.
      Store all standard solutions at -10°C to -20°C in screw-cap amber
      bottles with teflon liners.   Fresh standards  should be prepared every
      twelve months at a minimum.   The continuing calibration standard (50
      ng) should be prepared weekly and stored at 4°C (±2°C).

      In order to facilitate the confirmation of pesticides  and Aroclors from
      the semivolatile library search data (see Exhibit D PEST, paragraph
      17),  the laboratory may wish to include the pesticide/Aroclor target
      compounds listed in Exhibit  C in the semivolatile continuing
      calibration standard.   The laboratory may add any or all of these
      compounds to the semivolatile continuing calibration standard, but at a
      concentation of 10  ng/uL or  less.   If added to this GC/MS standard,
      these additional analytes are not reported on the semivolatile
      calibration form (Form VII),  but must be included in the quantitation
      report for the  continuing calibration standard.  As only a single point
      calibration would be  performed, no %RSD or percent difference criteria
      would apply to  these  additional analytes.
                                 D-38/SV                        OLM01.3   2/91

-------
                                                              SECTION IV


3.3   Instrument performance, check solution - .prepare a solution of
      decafluorotriphenylphosphine (DFTPP),  such that a 2 uL injection will
      contain 50 ng of DFTPP.   The DFTPP also  be included in the calibration
      standards at this level.

4.    Instrument Operating Conditions

      4.1     Gas Chroraatograph

              The following are the recommended GC analytical conditions:

              Initial  Column Temperature Hold   -       40°C for 4 minutes

              Column Temperature Program        -       40-270°C at 10
                                                        degrees/min.

              Final Column Temperature Hold     -       270°C for 10 minutes

              Injector Temperature              -       250-300°C

              Transfer Line Temperature         -       250-300°C

              Source Temperature                —       according to
                                                        manufacturer's
                                                        specifications

              Injector                         -       Grob-type, splitless

              Sample Volume                     -       2 uL

              Carrier  Gas                       -       Helium at 30 mL/sec

              Optimize GC conditions for analyte separation and sensitivity.
              Once optimized, the same GC conditions must be used for the
              analysis of all standards, samples, blanks, matrix spikes, and
              matrix spike duplicates.

      4.2     Mass Spectrometer

              The following are the required mass spectrometer analytical
              conditions:

              Electron Energy - 70 volts (nominal)

              Mass Range      - 35 to 500 amu

              Scan Time       - not to exceed 1 second per scan

      4.3     The GC/MS system must be tuned to meet the manufacturer's
              specifications, using a suitable calibrant such as FC-43 or
              perfluorokerosene (PFK).  The mass calibration and resolution
              of the GC/MS system are verified by the analysis of the
              instrument performance check solution  (paragraph 3.3).
                                 D-39/SV                       OLM01.2   1/91

-------
                                                              SECTION IV
               4.3.1      Prior to the analysis of any samples, blanks, or
                       -. calibration standards, the Contractor must establish
                         that the GC/MS system meets the mass spectral ion
                         abundance criteria for the instrument performance
                         check solution containing
                         decafluorotriphenylphosphine (DFTPP).

               4.3.2      The analysis of the instrument performance check
                         solution may be performed as follows:

                        o   As an injection of  up  to  50  ng  of DFTPP into  the
                            GC/MS
                        o   By adding  50 ng of  DFTPP  to  the calibration
                            standards  (paragraph 3.2) and analyzing the
                            calibration standard.
               4.3.3      The analysis of the instrument performance check
                         solution must meet the ion abundance criteria given
                         below.
                                   TABLE 1
 DFTPP KEY IONS AND ION ABUNDANCE CRITERIA FOR QUADRAPOLE MASS SPECTROMETERS
                     Ion Abundance Criteria

          51       30.0-80.0 percent of mass 198
          68       Less than 2.0 percent of mass 69
          69       Present
          70       Less than 2.0 percent of mass 69
         127       25.0-75.0 percent of mass 198
         197       Less than 1.0 percent of mass 198
         198       Base peak, 100 percent relative abundance (see note)
         199       5.0 - 9.0 percent of mass 198
         275       10.0 - 30.0 percent of mass 198
         365       Greater than 0.75 percent of mass 198
         441       Present but less than mass 443
         442       40.0 - 110.0 percent of mass 198
         443       15.0 - 24.0 percent of mass 442


NOTE:  All ion abundances MUST be normalized to m/z 198, the nominal base
peak, even though the ion abundances of m/z 442 may be up to 110 percent that
of m/z 198.
                                 D-40/SV                              OLM01.0

-------
                                                              SECTION IV
               4.3.4      The abundance criteria listed above.must be met for
                         a 50 ng injection of DFTPP.   The mass spectrum of
                         DFTPP must be acquired in the following manner.
                         Three scans (the peak apex scan and the scans
                         immediately preceding and following the apex) are
                         acquired and averaged.  .Background subtraction is
                         required, and must be accomplished using a single
                         scan prior to the elution of DFTPP.  Note:  All
                         subsequent standards, samples, MS/MSD, and blanks
                         associated with a DFTPP analysis must use identical
                         mass spectrometer instrument conditions.

               4.3.5      The criteria above are based on adherence to the
                         acquisition specifications identified in paragraph
                         4.3.4.  The criteria are based on performance
                         characteristics of instruments currently utilized in
                         routine support of Program activities.  These
                         specifications, in conjunction with relative
                         response factor criteria for 54 target compounds
                         (see Table 2), are designed to control and monitor
                         instrument performance associated with the
                         requirements of this Statement of Work.

               4.3.6      The instrument performance check solution must be
                         analyzed once at the beginning of each 12-hour
                         period during which samples or standards are
                         analyzed.

                         The twelve (12) hour time period for a GC/MS system
                         instrument performance check and standards
                         calibration (initial or continuing calibration
                         criteria) begins at the moment of injection of the
                         DFTPP analysis that the laboratory submits as
                         documentation of a compliant instrument performance
                         check.  The time period ends after twelve -(12) hours
                         has elapsed according to the system clock.

5.     Calibration

5.1   Prior to the analysis of samples and required blanks,  and after the
      instrument performance check solution criteria have been met, each
      GC/MS system must be calibrated at a minimum of five concentrations to
      determine instrument sensitivity and the linearity of GC/MS response
      for the semivolatile target compounds.

5.2   The internal standards are added to all calibration standards and all
      sample extracts (including blanks, matrix spikes, and matrix spike
      duplicates) just prior to analysis by GC/MS.  A 10 uL aliquot of the
      internal standard solution should be added to a 1 mL aliquot of
      calibration standards.  The internal standards specified in paragraph
      3.1 should permit most of the semivolatile target compounds to have
      relative retention times of 0.80 to 1.20, using the assignments of
      internal standards to target compounds given in Table 2.
                                  D-41/SV                        OLM01.3  2/91

-------
                                                              SECTION IV
5.3   The. quantitation ions for each internal standard are given in Table 3.
      Use the primary ion listed in Table 3 for quantitation,  unless
      interferences are present.  If interferences prevent the use of the
      primary ion for a given internal standard, use the secondary ion(s)
      listed in Table 3.

5.4   Prepare calibration standards at a minimum of five concentration levels
      for each target compound and surrogate, as specified in paragraph 3.2.
      Analyze 2 uL of each calibration standard and tabulate the area of the
      primary characteristic ion against concentration for each compound
      including the surrogate compounds.  A 2 uL injection is required.
      Calculate relative response factors (RRF) for each compound using
      Equation 1.
          RRF -
                    Ax      cis
      Where,

      Ax  -   Area of  the characteristic ion for the compound  to be measured
              (see Table 4)

      A^s =   Area of  the characteristic ion for the specific  internal
              standard (see Table 3)

      cis =   Concentration of  the  internal standard (ng/uL)

      Cx  -   Concentration of  the  compound to be measured  (ng/uL)


5.5   The average relative response factor (RRF)  must be calculated for all
      compounds.  Calculate the % Relative Standard Deviation (%RSD) of the
      RRF values for the initial calibration using the following equation:

              %RSD = Standard Deviation  x 100
                         Mean
              Where,
              Standard Deviation
                                        n
                                            ci - x)'
                                             n-1
1/2
              Where

              Xi = each individual value used to calculate the mean

              x  = the mean of n values

              n  = the total number  of values



                                 D-42/SV                        OLM01.2  1/91

-------
     SEMIVOLATILE INTERNAL STANDARDS WIT!!
                                                             TABLE  2
                                                :ORRESPONDING TARGET COMPOUNDS AND SURROGATES ASSIGNED FOR QUANTITATION
1,4-Dichlorobenzene-d^  Naphthalene-dg     Acenaphthene-d-, Q      Phenanthrene-d,Q   Chrysene-di,       Perylene-d^
o

o
I—>
o
Phenol
bis(2-Chloroechyl)
 ether
2 -Chlorophenol
1,3-Dichlorobenzene
1,4-Dichlorobenzene
1,2 -Dichlorobenzene
2-Methylphenol
2,2'-oxybis-
 (1-Chloropropane)
4-Methylphenol
N-Nitroso-Di-n-
 propylainine
Hexachloroethane
2 -Fluorophenol
 (surr)
Phenol-d^  (surr)
2-Chlorobenzene-d,
 (surr)
1, 2 -Dichlorobenzene-d,
 (s u r r)
                              Nitrobenzene
                              Isophorone
                              2-Nitrophenol
                              2,4-Dimethyl-
                               phenol
                              bis(2-Chloro-
                                e thoxy)me thane
                              2,4-Dichloro-
                               phenol
                              1,2,4-Trichloro-
                               benzene
                              Naphthalene
                              4-Chloroaniline
                              Hexachloro-
                               butadiene
                              4-Chloro-3-
                                methylphenol
                              2-Methylnaphth-
                               alene
                              Nitrobenzene-de
                               (surr)
Hexachlorocyclo-
 pentadiene
2,4,6-Trichloro-
 phenol
2,4,5-Trichloro-
 phenol
2 -Chloronaphthalene
2-Nitroaniline
Dimethyl Phthalate
Acenaphthylene
3-Nitroaniline
Acenaphthene
2,4-Dinitrophenol
4-Nitrophenol
Dibenzofuran
2 ,4-Dinitrotoluene
2,6-Dinitrotoluene
Diethyl phthalate
4 -Chlorophenyl
 phenyl ether
Fluorene
4-Nitroaniline
2 -Fluorob iphenyl
 (surr)
2,4,6-Tribromo-
 phenol (surr)
4,6-Dinitro-2-
 methylphenol
;,-nitrosodi-
 phenylamine
4-Bromophenyl
 phenyl ether
Hexachloro-
 benzene
Pentachloro-
 phenol
Phenanthrene
Carbazole
Anthracene
Di-n-butyl-
 phthalate
Fluoranthene
Pyrene
Butylbenzyl
 phthalate
3,3'-Dichloro-
 benzidine
 Benzo(a) -
 anthracene
bis(2-Ethyl-
 hexyl)phthalate
Chrysene
 (surr)
Di-n-octyl-
 phthalate
Benzo(b)fluor-
 anthene
Benzo(k)fluor-
 anthene
Benzo(a)pyrene
Indeno(l,2,3-cd)
 pyrene
Dibenz(a.h)-
 anthracene
Benzo(g,h,i)-
 perylene
                                                                                                                  in
                                                                                                                  m
                                                                                                                  o
                                                                                                                  H
                                                                                                                  M
                                                                                                                  O
                                                                                                                  •z
surr = surrogate compound

-------
                                                             SECTION IV
                                    TABLE  3
        CHARACTERISTIC  IONS  FOR INTERNAL STANDARDS  FOR  SEMIVOLATILE  COMPOUNDS
INTERNAL STANDARDS                Primary Ion             Secondary Ions


l,4-Dichlorobenzene-d4                152                    115
Naphthalene-ds                        136                     68
Acenapthene-diQ                       164                    162, 160
Phenanthrene-d^Q                      188                     94,  80

Chrysene-di2                          24°                    12°- 236
Perylene-d12                          264                    260, 265
                                D-44/SV                              OLM01.0

-------
                                                             SECTION IV
                                   TABLE 4
     CHARACTERISTIC IONS FOR SEMIVOLATILE TARGET COMPOUNDS AND SURROGATES
Parameter
Phenol
bis(2-Chloroethyl)ether
2 - Ch 1 o r opheno 1
1 , 3 - Dichlorobenzene
1 , 4 -Dichlorobenzene
1,2- Dichlorobenzene
2-Methylphenol
2,2' -oxybis(l-Chloropropane)
4-Methylphenol
N-Nitroso-di-propylamine
Hexachloroethane
Nitrobenzene
Isophorone
2-Nitrophenol
2 , 4 - Dime thylphenol
bis(2-Chloroethoxy)methane
2 ,4-Dichlorophenol
1,2 ,4-Trichlorobenzene
Naphthalene
4-Chloroaniline
Hexachlorobutadiene
4 -Chloro- 3 -me thylphenol
2-Methylnaphthalene
Hexachlorocyclopentadiene
2,4, 6 -Trichlorophenol
2,4, 5 -Trichlorophenol
2 - Chloronaphthalene
2-Nitroaniline
Dimethyl phthalate
Acenaphthylene
3-Nitroaniline
Acenaphthene
2 ,4-Dinitrophenol
4-Nitrophenol
Dibenzofuran
2 ,4-Dinitrotoluene
2 , 6-Dinitrotoluene
Primary .Ion
94
93
128
146
146
146
108
45
108
70
117
77
82
139
107
93
162
180
128
127
225
107
142
237
196
196
162
65
163
152
138
153
184
109
168
165
165
Secondary
65,
63,
64,
148,
148,
148,
107
77,
107
42,
201,
123,
95,
65,
121,
Ion(s)
66
95
130
113
113
113

79

101, 130
199
65
138
109
122
95, 123
164,
182,
129,
129
223,
144,
141
235,
198,
198,
164,
92,
194,
151,
108,
152,
63,
139,
139
63,
89,
98
145
127

227
142

272
200
200
127
138
164
153
92
154
154
65

182
121
(continued)
                                D-45/SV
OLM01.0

-------
                                                             SECTION IV
                              TABLE  4  (continued)
     CHARACTERISTIC IONS FOR SEMIVOLATILE TARGET COMPOUNDS AND SURROGATES
Parameter
Primary Ion    Secondary Ion(s)
Diethylphthalate
4-Chlorophenyl-phenylether
Fluorene
4-Nitroaniline
4,6-Dinitro-2-methylphenol
N-Nitrosodiphenylamine
4-Bromophenyl-phenylether
Hexachlorobenzene
Pentachlorophenol
Phenanthrene
Anthracene
Carbazole
Di-n-butylphthalate
Fluoranthene
Pyrene
Butylbenzylphthalate
3,3'-Dichlorobenzidine
Benzo(a)anthracene
bis(2 -Ethylhexyl)phthalate
Chrysene
Di-n-Octyl phthalate
Benzo(b)fluoranthene
Benzo(k)fluoranthene
Benzo(a)pyrene
Indeno(l,2,3-cd)pyrene
Dibenz(a,h)anthracene
Benzo(g,h,i)perylene

SURROGATES
   149
   204
   166
   138
   198
   169
   248
   284
   266
   178
   178
   167
   149
   202
   202
   149
   252
   228
   149
   228
   149
   252
   252
   252
   276
   278
   276
177, 150
206, 141
165, 167
 92, 108
182,  77
168, 167
250, 141
142, 249
264, 268
179, 176
179, 176
166, 139
150, 104
101, 100
101, 100
 91, 206
254, 126
229, 226
167, 279
226, 229

253, 125
253, 125
253, 125
138, 227
139, 279
138, 277
Phenol -d5
2 - Fluorophenol
2 ,4,6-Tribromophenol
Nitrobenzene -d$
2-Fluorobiphenyl
Terphenyl
2-Chlorophenol-d4
1,2- Dichlorobenzene - d^
99
112
330
82
172
244
132
152
42,
64
332,
128,
171
122,
68,
115,
71

141
54

212
134
150
                                D-46/SV
                            OLM01.0

-------
                                                             SECTION IV
5.6   Response factor criteria have been established for the calibration of
      the semivolatile target compounds and semivolatile surrogate compounds.

      5.6.1   The response factors of  the compounds listed  in Table 5 must
              meet the minimum RRF criteria at each concentration level and
              maximum %RSD criteria for the initial calibration, with
              allowance made for up to four semivolatile target and surrogate
              compounds.  However, the RRFs for those four  compounds must be
              greater than 0.010, and  the %RSD of those four compounds must
              be less than or equal to 40.0% for the initial calibration to
              be acceptable.

                                   TABLE 5
         RELATIVE RESPONSE FACTOR CRITERIA FOR INITIAL AND CONTINUING
                 CALIBRATION OF SEMIVOLATILE TARGET COMPOUNDS
Semivolatile
Compounds
Phenol
bis(-2-Chloroethyl)ether
2-Chlorophenol
1 , 3-Dichlorobenzene
1 ,4-Dichlorobenzene
1 , 2-Dichlorobenzene
2-Methylphenol
4-Methylphenol
N-Nitroso-Di-propylamine
Hexachloroe thane
Nitrobenzene
Isophorone
2-Nitrophenol
2 ,4-Dimethylphenol
bis(-2-Chloroethoxy)methane
2,4-Dichlorophenol
1 , 2 ,4-Trichlorobenzene
Naphthalene
4-Chloro-3-methylphenol
2 -Methylnaphthalene
2 ,4,6-Trichlorophenol
2,4, 5 -Trichlorophenol
2 - Chloronaphthalene
Acenaphthylene
2 , 6-Dinitrotoluene
Acenaphthene
Dibenzofuran
2 ,4-Dinitrotoluene
4 - Chlorophenyl -phenyle ther
Fluorene
4 -Bromophenyl- phenyle ther
Hexachlorobenzene
Pentachlorophenol
Minimum
RRF
0.800
0.700
0.800
0.600
0.500
0.400
0.700
0.600
0.500
0.300
0.200
0.400
0.100
0.200
0.300 '
0.200
0.200
0.700
0.200
0.400
0.200
0.200
0.800
1.300
0.200
0.800
0.800
0.200
0.400
0.900
0.100
0.100
0.050
Maximum
%RSD
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
Maximum
%Diff
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
(continued)
                                D-47/SV
OLM01.1  12/90

-------
                                                             SECTION IV
                             TABLE  5  (continued)
         RELATIVE RESPONSE FACTOR CRITERIA FOR INITIAL AND CONTINUING
                 CALIBRATION OF SEMIVOLATILE TARGET COMPOUNDS
Semivolatile
Compounds
Phenanthrene
Anthracene
Fluoranthene
Pyrene
Benzo ( a) anthracene
Chrysene
Benzo (b)fluoranthene
Benzo (k) f luoranthene
Benzo(a)pyrene
Indeno(l,2, 3-cd)pyrene
Dibenzo (a , h) anthracene
Benzo(g,h, i)perylene
Nitrobenzene -d$
2 - Fluor obiphenyl
Terphenyl-di4
Phenol -d5
2 - Fluorophenol
2 - Chlorophenol - d4
1 , 2 - Dichlorobenzene - d4
Minimum
RRF
0.700
0.700
0.600
0.600
0.800
0.700
0.700
0.700
0.700
0.500
0.400
0.500
0.200
0.700
0.500
0.800
0.600
0.800
0.400
Maximum
%RSD
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
Maximum
%Diff
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
      5.6.2
         The  following compounds have no Maximum  %RSD, or Maximum
         %Difference criteria; however, these compounds must meet  a
         minimum RRF criterion of  0.010:
              2,2'-oxybis(l-Chloropropane)
              4-Chloroaniline
              Hexachlorobutadiene
              Hexachlorocyclopentadiene
              2-Nitroaniline
              DimethyIphthalate
              3-Nitroaniline
              2,4-Dinitrophenol
              4-Nitrophenol
              Diethylphthalate
                                              4-Nitroaniline
                                              4,6-Dinitro-2-methylphenol
                                              N-Nitrosodiphenylamine
                                              Di-n-butylphthalate
                                              Butylbenzylphthalate
                                              3,3'-Dichlorobenzidine
                                              bis(2:Ethylhexyl)phthalate
                                              Di-n-octyIphthalate
                                              2,4,6-Tribromophenol
                                              Carbazole
5.7
A check of the calibration curve must be performed once every 12 hours
(see paragraph 4.3.6 for the definition of the twelve-hour time period),
Check the relative response factors of those compounds for which RRF
values have been established.  If these criteria are met, the relative
response factors for all compounds are calculated and reported.  A
percent difference of the daily relative response factor (12 hour)
compared to the average relative response factor from the initial curve
is calculated. Calculate the percent difference for each compound and
compare with the maximum percent difference criteria listed above.  For
negative percent difference values, the value must be greater than or
equal to -25.0%, but less than 0%.
                                D-48/SV
                                                         OLM01.2  1/91

-------
                                                             SECTION IV
      As with the initial calibration, up to four semivolatile target
      compounds in Table 5 may fail to meet the minimum RRF or maximum %D
      criteria, but the RRFs of those four compounds must be greater than or
      equal to 0.010, and the percent differences must be less than or equal
      to 40.0% for the continuing calibration to be acceptable.

5.8   Internal standard responses and retention times in all standards must be
      evaluated during or immediately after data acquisition. If the retention
      time for any internal standard changes by more than 0.50 minutes (30
      seconds) from the latest daily (12 hour) calibration standard, the
      chromatographic system must be inspected for malfunctions, and
      corrections made as required.  The extracted ion current profile (EICP)
      of the internal standards must be monitored and evaluated for each
      standard.  If the EICP area for any internal standard changes by more
      than a factor of two (-50% to +100%), the mass spectrometric system must
      be inspected for malfunction, and corrections made as appropriate.  When
      corrections are made, re-analysis of samples analyzed while the system
      was malfunctioning is necessary.

5.9   Each GC/MS system must be calibrated upon award of the contract,
      whenever the Contractor takes corrective action which may change or
      affect the initial calibration criteria (i.e., ion source cleaning or
      repair, column removal or replacement, etc.), or if the continuing
      calibration acceptance criteria have not been met.

5.10  If time remains in the 12 hour time period after meeting the acceptance
      criteria for the initial calibration, samples may be analyzed.  It is
      not necessary to analyze a continuing calibration standard, if the
      initial calibration standard that is the same concentration as the
      continuing calibration standard meets the continuing calibration
      acceptance criteria.   Quantify all sample results against the initial
      calibration standard that is the same concentration as the continuing
      calibration standard (50ng/2uL).

5.11  If time does NOT remain in the 12-hour period beginning with the
      injection of the instrument performance check solution, a new injection
      of the instrument performance check solution must be made.  The DFTPP
      may be included in the continuing calibration standard.

5.12  If the injection of the instrument performance check solution meets the
      criteria in Table 1,  calculate the response factors for the continuing
      calibration standard and the percent difference of the response factors
      from the mean response factors in the initial calibration.

5.13  The response factors  from the continuing calibration standard must meet
      the criteria in Table 5 prior to the analysis of any blanks or samples.
                                D-49/SV                              OLM01.0

-------
                                                             SECTION IV

6.     Sample Analysis

6.1   Sample extracts may be analyzed only after the GC/MS system has met the
      instrument performance check,  initial calibration,  and continuing
      calibration requirements above.   The same instrument conditions must be
      employed for the analysis of samples as were used for calibration.

6.2   Internal standard solution is added to each sample extract.  Add 10 uL
      of internal standard solution to each accurately measured 1.0 mL of
      water sample extract.  For soil samples and water samples subjected to
      GPC, add 5 uL of internal standard solution to each accurately measured
      0.5 mL of sample extract.  This will result in a concentration of  20
      ng/uL of each internal standard.

6.3   Make any extract dilution indicated by.characterization prior to the
      addition of internal standards.   If any further dilutions of water or
      soil/sediment extracts are made, additional internal standards must be
      added to maintain the required 40 ng (20 ng/uL) of each internal
      standard in the extract volume.

6.4   Inject 2 uL of the sample extract into the GC/MS.  This 2 uL volume must
      contain 40 ng of each internal standard.

7.     Qualitative Analysis

7.1   The compounds listed in the Target Compound List (TCL), Exhibit C,  shall
      be identified by an analyst competent in the interpretation of mass
      spectra (see Exhibit A, Section III) by comparison of the sample mass
      spectrum to the mass spectrum of a standard of the suspected compound.
      Two criteria must be satisfied to verify the identifications:
      o  Elution  of  the  sample  component at  the GC relative  retention  time as
         the standard  component.
      o  Correspondence  of  the  sample  component and  standard component mass
         spectra.

      7.1.1   For establishing correspondence of  the GC  relative  retention
              time  (RRT),  the  sample  component RRT  must  compare  within +0.06
              RRT units  of the RRT of the standard  component.  For  samples
              analyzed  during  the same  12-hour time  period  as  the  initial
              calibration  standards,  compare  the  sample  retention  times  to
              those from the 50 ng calibration standard.  For  reference,  the
              standard  must be run on the same shift as  the sample.   If
              coelution of interfering  components prohibits accurate
              assignment of  the sample  component  RRT  from the  total  ion
              chromatogram,  the RRT should  be assigned by using  extracted  ion
              current profiles for  ions unique to the  component  of  interest.
      7.1.2   For comparison of standard  and  sample  component  mass  spectra,
              mass  spectra obtained on  the  contractor's  GC/MS  are  required.
              Once  obtained, these  standard spectra  may  be  used  for
              identification purposes,  only if the  contractor's  GC/MS meets
              the DFTPP daily  instrument  performance  requirements.   These
              standard  spectra may be obtained from the  run used to  obtain
              reference RRTs.

      7.1.3   The requirements for  qualitative verification by comparison  of
              mass  spectra are as  follows:

                                 D-50/SV                       OLM01.3  2/91

-------
                                                             SECTION IV
               7.1.3.1    All ions present in the standard mass spectra at a
                         relative intensity greater than 10% (most abundant
                         ion in the spectrum equals 100%) must be present in
                         the sample spectrum.

               7.1.3.2    The relative intensities of ions specified in
                         7.1.3.1 must agree within +20% between the standard
                         and sample spectra.  (Example:  For an-ion with an
                         abundance of 50% in the standard spectra, the
                         corresponding sample ion abundance must be between
                         30 and 70 percent.)

               7.1.3.3    Ions greater than 10% in the sample spectrum but not
                         present in the standard spectrum must be considered
                         and accounted for by the analyst making the
                         comparison. In Task III, the verification process
                         should favor false positives.   All compounds meeting
                         the identification criteria must be reported with
                         their spectra.  For all compounds below the CRQL
                         report the actual value followed by "J", e.g., "3J."

      7.1.4    If a compound cannot be verified by all of the criteria  in
               7.1.3, but in the technical judgement of the mass spectral
               interpretation specialist, the identification is correct, then
               the Contractor shall report that identification and proceed
               with quantification in paragraph 8.

7.2   A library search shall be executed for non-target sample components for
      the purpose of tentative identification.   For this  purpose,  the most
      recent release of the NIST/EPA/MSDC mass  spectral library,  shall be
      used.

      7.2.1    Up to 20 nonsurrogate organic compounds of greatest apparent
               concentration not listed in Exhibit C for the semivolatile
               fraction shall be identified tentatively via a forward  search
               of the NIST/EPA/MSDC mass spectral library.  Substances  with
               responses less than 10% of the nearest internal standard are
               not required to be searched in this fashion.  Only after visual
               comparison of sample spectra with  the nearest library searches
               will the mass spectral interpretation specialist assign  a
               tentative identification.  NOTE:   Computer generated library
               search routines must not use normalization routines that would
               misrepresent the library or unknown spectra when compared to
               each other.

               Peaks that are suspected as aldol-condensation reaction
               products (i.e., 4-methyl-4-hydroxy-2-pentanone and 4-methyl-3-
               pentene-2-one) shall be searched and reported but not counted
               as part of the 20 most intense non-target semivolatile
               compounds.

      7.2.2    Guidelines for making tentative identification.

               7.2.2.1    Relative intensities of major ions in the reference
                         spectrum (ions greater  than 10% of the most  abundant
                         ion) should be present  in the sample spectrum.

                                D-51/SV                       OLM01.3  2/91

-------
                                                             SECTION IV
               7.2.2.2    The relative intensities of the major ions should
                         agree within + 20%.  (Example:   For an ion with an
                         abundance of 50% in the standard spectra,  the
                         corresponding sample ion abundance must be between
                         30 and 70 percent.)

               7.2.2.3    Molecular ions present in reference spectrum should
                         be present in sample spectrum.

               7.2.2.4    Ions present in the sample spectrum but not in the
                         reference spectrum should be reviewed for possible
                         background contamination or presence of co-eluting
                         compounds.

               7.2.2.5    Ions present in the reference spectrum but not in
                         the sample spectrum should be reviewed for possible
                         subtraction from the sample spectrum because of
                         background contamination or coeluting compounds.
                         NOTE:  Data system library reduction programs
                         sometimes can create these discrepancies.

      7.2.3    If, in the technical judgement of the mass interpretation
               spectral specialist, no valid tentative identification can be
               made, the compound should be reported as unknown.  The mass
               spectral specialist should give additional classification of
               the unknown compound, if possible (i.e., unknown phthalate,
               unknown hydrocarbon, unknown acid type, unknown chlorinated
               compound). If probable molecular weights can be distinguished,
               include them.

8.     Ouantitation

8.1   Target components  identified shall be  quantified by the internal
      standard method.   The internal standard used shall be the one nearest
      the retention time to that of  a given  analyte (see Table 2).   The EICP
      area of characteristic ions of analytes listed in  Table 4 are used for
      quantitation.   In  all instances where  the data system report has been
      edited, or where manual integration or quantitation has been performed,
      the GC/MS operator must identify such  edits  or manual procedures by
      initializing and dating the changes made to  the report.

      Internal standard  responses and retention times in all samples must be
      evaluated during or immediately after  data acquisition.  If the
      retention time for any internal standard changes by more than 0.50
      minutes (30 seconds)  from the  latest daily (12 hour)  calibration
      standard, the chromatographic  system must be inspected for malfunctions,
      and corrections made  as required.  For  samples analyzed during the same
      12-hour time period as the initial calibration standards,  compare the
      internal standard  responses and retention times to those of the 50 ng
      calibration standard.   The extracted ion current profile (EICP) of the
      internal standards  must be monitored and evaluated for each sample,
      blank,  matrix spike,  and matrix spike  duplicate.  The criteria are
      described in detail in the instructions for Form VIII, Internal Standard
      Area Summary.   If  the EICP area for any internal standard changes by
      more than a factor  of two (-50% to +100%),  the mass spectrometric system
      must be inspected  for malfunction and  corrections  made as appropriate.

                                D-52/SV                       OLM01.2   1/91

-------
                                                             SECTION IV
      If the analysis of a subsequent sample or standard indicates that the
      system is functioning properly, then corrections may not be required.
      The samples or standards with EICP areas outside the limits must be
      re-analyzed, and treated according to paragraphs 8.1.1 and 8.1.2 below.
      If corrections are made, then the laboratory must demonstrate that the
      mass spectrometric system is functioning properly.  This must be
      accomplished by the analysis of a standard or sample that does meet the
      EICP criteria.  After corrections are made,  the re-analysis of samples
      analyzed while the system was malfunctioning is required.

      8.1.1    If  after re-analysis,  the EICP areas for  all internal standards
               are inside the  contract  limits  (-50% to +100%),  then the
               problem with the first analysis  is considered to have been
               within the control  of  the laboratory.  Therefore, submit only
               data  from the analysis with EICPs within  the contract limits.
               This  is considered  the initial analysis and must be reported as
               such  on all data deliverables.

      8.1.2    If  the re-analysis  of  the sample does not solve  the problem,
               i.e., the EICP  areas are outside the contract limits for both
               analyses, then  submit  the EICP data and sample data from both
               analyses.  Distinguish between the initial analysis and the
               re-analysis on  all  data  deliverables, using the  sample suffixes
               specified in Exhibit B.  Document in the  SDG Narrative all
               inspection and  corrective actions taken.

      8.1.3    Do  not re-analyze MS/MSD samples that do  not meet the EICP area
               limits.

8.2   The relative response factor (RRF) from the daily standard analysis is
      used to calculate the concentration in the sample.  For samples analyzed
      during the same 12-hour time period as the initial calibration
      standards, use the RRF values from the 50 ng calibration standard.
      Secondary ion quantitation is allowed ONLY when there are sample
      interferences with the primary ion.   If secondary ion quantitation is
      performed, document the reasons in the SDG Narrative.   The area of a
      secondary ion cannot be substituted for the area of a primary ion unless
      a relative response factor is calculated using the secondary ion.

      When target compounds are below contract required quantitation limits
      (CRQL) but the spectrum meets the identification criteria, report the
      concentration with a "J."  For example,  if CRQL is 10 ug/L and
      concentration of 3 ug/L is calculated, report as "3J."

      Calculate the concentration in the sample using the relative response
      factor (RRF) as determined in 5.4 and the following equation:
                                           (Ax)(Is)(Vt)(Df)
                               ug/L  =  	
                                        (Ais)(RRF)(V0)(Vi)
                                D-53/SV                        OLM01.2  1/91

-------
                                                              SECTION IV


               Where,

               Ax  =•  Area of the characteristic ion for the compound to be
                      measured

               A^g =  Area of the characteristic ion for the internal standard

               Is  -  Amount of internal standard injected in nanograms (ng)

               Vo  =  Volume of water extracted in milliliters (mL)

               V^  =  Volume of extract injected in microliters  (uL)

               Vt  =  Volume of the concentrated extract in microliters (uL)

               Df  =  Dilution Factor.   The dilution factor for  analysis of
                      water samples for semivolatiles by this method is
                      defined as follows:

               uL most cone,  extract used  to make dilution + uL  clean solvent
                         uL most cone,  extract  used to make dilution

               If no dilution is performed,  Df  = 1.0.


               Soil/Sediment

                                             (Ax)(Is)(Vt)(Df)(2.0)
               Concentration     ug/kg  =
               (Dry weight basis)          (Ais)(RRF)(V^(Ws)(D)

               Where,

               AX,IS,A^S  are as  given for water,  above.

               Vt  =  Volume of the concentrated extract in microliters (uL)

               Vi  =  Volume of extract injected in microliters (uL)

               D    =  100 - % moisture
                          100

               Ws  =  Weight of sample extracted in grams (g)


               Df  =  Dilution Factor.  The dilution factor for analysis of
                      soil samples for semivolatiles by this method is defined
                      as  follows:

               uL most cone,  extract  used to make dilution  + mL clean solvent
                       uL most  cone,  extract used to make  dilution

               If no dilution is performed, Df =  1.0.

The factor of 2.0 in the numerator is used to account for the amount of
extract that is not recovered from the mandatory use of GPC cleanup.
Concentrating the extract collected


                                  D-54/SV                       OLM01.3  2/91

-------
                                                             SECTION IV
              after GPC to 0.5 mL, rather than 1.0 mL for water samples not
              subjected to GPC, maintains the sensitivity of the soil method
              comparable to that of the water method, but correction of the
              numerical result is still required.

8.3   An estimated concentration for non-target components tentatively
      identified shall be quantified by the  internal standard method.   For
      quantification,  the nearest internal standard free of interferences
      shall be used.   The formula for calculating concentrations is the same
      as in paragraph 8.2.   Total area counts (or peak heights) from the total
      ion chromatograms are to be used for both the compound to be measured
      and the internal standard.   A relative response factor (RRF) of one  (1)
      is to be assumed.  The resulting concentration shall be qualified as "J"
      (estimated, due to lack of a compound-specific response factor),  and "N"
      (presumptive evidence of presence),  indicating the quantitative and
      qualitative uncertainties associated with this non-target component.  An
      estimated concentration should be calculated for all tentatively
      identified compounds as well as those  identified as unknowns.

8.4   If the on-column concentration of any  compound in any sample exceeds the
      initial calibration range,  that sample extract must be diluted,  the
      internal standard concentration readjusted, and the sample extract
      reanalyzed.  Guidance in performing dilutions and exceptions to this
      requirement are given below.

      8.4.1   Use the  results of the original analysis  to determine the
              approximate dilution factor required  to get the  largest  analyte
              peak within the initial calibration range.

      8.4.2   The dilution factor chosen should  keep the response  of the
              largest  peak for a target compound in the upper  half of  the
              initial  calibration range of  the instrument.

      8.4.3   Do not submit data for more than two analyses, i.e., the
              original sample extract and one dilution, or,  if the
              semivolatile screening procedure was employed, from  the  most
              concentrated dilution analyzed and one further dilution.

      8.4.4   Do not dilute MS/MSD samples  to get either spiked or non-spiked
              analytes within the calibration range. If the  sample from which
              the MS/MSD aliquots were  taken contains high  levels  of the  .
              spiked analytes, calculate the concentration  and recovery of
              the analytes from the undiluted analysis,  and  note  the problem
              in the SDG Narrative.

8.5   Calculate surrogate standard recovery on all samples, blanks, and
      spikes.  Determine if recovery is within limits (see Table 6) and report
      on appropriate form.

      8.5.1   Calculate the concentrations  of the surrogate  compounds  using
              the same equations as used for the target compounds.  Calculate
              the recovery of each surrogate as  follows:

              %Recovery = Concentration (or amount) found   x 100
                          Concentration (or amount) spiked
                                 D-55/SV                        OLM01.2  1/91

-------
                                                       SECTION IV
8.5.2   Determine  if  the sample surrogate recovery meets specifications
        as  follows:

        o   The eight semivolatile surrogates can be divided into three
            groups:  base/neutral compounds (Nitrobenzene-d5,  2-
            Fluorobiphenyl, and Terphenyl-d]^);   acid compounds (Phenol-
            ds, 2-Fluorophenol, and 2,4,6-fribromophenol);   and
            compounds with advisory QC limits (2-Chlorophenol-d4 and
            1,2-Dichlorobenzene-d^.

        o   If a single surrogate recovery from any group is not within
            the contract windows,  the sample does not require reanalysis
            or re-extraction.

        o   If a single surrogate recovery from the base/neutral group
            and a single surrogate recovery from the acid group are not
            within the contract windows,  the sample does not require
            reanalysis or re-extraction.

        o   Do not reanalyze or re-extract if only surrogates with
            advisory QC limits are not within the contract windows.

8.5.3   If  the sample surrogate recovery does not meet specifications
        (i.e., if  two base/neutral or  two acid surrogates are out  of
        limits or  if  recovery  of any one base/neutral or acid surrogate
        is  below 10%),  the  following are required:

        o   Check to be sure that there are no errors in calculations,
            surrogate solutions, and internal standards.   Also check
            instrument performance.
        o   Reanalyze the sample if none  of the above reveal a problem.
        o   If surrogate recoveries in *a  blank do not meet
            specifications, the blank may be reanalyzed alone.

        o   Do not reanalyze dilutions if surrogate recoveries are
            outside the limits.

        o   Never reanalyze the matrix spike or matrix spike duplicate
            (MS/MSD),  even if surrogate recoveries are outside the
            limits.

        o   If the sample associated with the matrix spike and matrix
            spike duplicate does not meet specifications, it should be
            reanalyzed only if the MS/MSD surrogate recoveries are
            within the limits.  If the sample and associated MS/MSD show
            the same pattern (i.e.,  outside the limits),  then the sample
            does not require reanalysis and a re-analysis must not be
            submitted.

            Document in the narrative the similarity in surrogate
            recoveries.

8.5.4   If  the reanalysis  of the sample solves the problem,  then the
        problem was within the laboratory's control.  Therefore,  submit
        only data  from  the  analysis with surrogate spike recoveries
        within the contract windows.   This shall be  considered the
        initial analysis and shall be  reported as such on all  data
        deliverables.


                           D-56/SV                        OLM01.2  1/91

-------
                                                       SECTION IV
8.5.5    If none  of  the  steps  in paragraph 8.5.3 or 8.5.4 solves  the
         problem,  then,  except as noted below, re-extract and  reanalyze
         the sample.   If the re-extraction and reanalysis of the  sample
         solves the  problem, then the problem was within the
         laboratory's  control.  Therefore, submit only data from  the
         analysis with surrogate recoveries within the contract windows.
         This shall  be considered the initial analysis and shall  be
         reported as such on all data deliverables.

         o    If surrogate recoveries in a blank do not meet
             specifications even after reanalysis,  all of the samples
             associated with that blank must be re-extracted along with
             the blank.  The blank is intended to detect contamination in
             samples processed at the same time.
         o    Do not re-extract diluted samples if surrogate recoveries
             are outside the limits.
         o    Never re-extract the matrix spike or matrix spike duplicate
             (MS/MSD),  even if surrogate recoveries are outside the
             limits.

         o    If the sample associated with the matrix spike and matrix
             spike duplicate does not meet specifications after
             reanalysis,  it should be reextracted only if the reanalysis
             surrogate recoveries are not within  the  limits and MS/MSD
             surrogate recoveries are within the  limits.   If the sample
             and associated MS/MSD show the same  pattern (i.e., outside
             the limits), then the sample does not require reanalysis and
             a reanalysis must not be submitted.
            Document in the narrative the similarity in surrogate
            recoveries.

8.5.6    If the re-extraction  and reanalysis of the sample does not
         solve the problem  (i.e., the surrogate recoveries are outside
         the  contract  limits for both analyses), then submit the
         surrogate recovery data and sample analysis data from' the
         initial  analysis of both sample extracts  (e.g., the first
         analysis  of both extracts  of the sample).  Distinguish between
         the  initial analysis  and the analysis of the re-extracted
         sample on all data .deliverables, using the sample suffixes
         supplied in Exhibit B.
                          D-57/SV                        OLM01.2  1/91

-------
                                                             SECTION IV
                                   TABLE 6
                          SURROGATE RECOVERY LIMITS
Compound
Nitrobenzene -d5
2-Fluorobiphenyl
Terphenyl - d^4
Phenol -d5
2-Fluorophenol
2,4, 6-Tribromophenol
2 - Chlorophenol - d4
1 , 2 - Dichlorobenzene - d4
%Recovery
Water
35-114
43-116
33-141
10-110
21-110
10-123
33-110
16-110
%Recovery
Soil
23-120
30-115
18-137
24-113
25-121
19-122
20-130
20-130






(advisory)
(advisory)
8.6.   A matrix spike and matrix spike duplicate must be performed  for each
      group of samples of a similar matrix,  for the following,  whichever is
      most frequent:

            o   Each Case of  field  samples received, OR

            o   Each 20  field samples  in a Case, OR

            o   Each group  of field samples of  a similar concentration level
                (soils only) , OR

            o   Each 14  calendar day period (7  calendar day period  for 14 -day
                data turnaround contracts) during which field samples  in a
                Case were received  (said period beginning with  the  receipt  of
                the  first sample in that Sample Delivery Group) .
      Calculate the  recovery of each matrix  spike  compound in the matrix spike
      and matrix spike duplicate  and report  on appropriate form.

      8.6.1.   Calculate the  concentrations of  the matrix spike compounds
              using  the same equations as used for target compounds.
              Calculate the  recovery of each matrix spike compound as
              follows:

                                      SSR  -  SR
            Matrix Spike Recovery = - x 100
              Where,

              SSR  = Spike sample result
              SR   = Sample result
              SA   = Spike added

      8.6.2    Calculate the relative percent difference of the recoveries of
              each compound in the matrix spike and matrix spike duplicate as
              follows :

                      |MSR - MSDR|
              RPD = _ x 100
                     (1/2)(MSR+MSDR)

                                D-58/SV                       OLM01.2   1/91

-------
                                                             SECTION IV
      8.6.3
    Compound
Where,

RPD = Relative Percent Difference

MSR = Matrix Spike Recovery

MSDR = Matrix Spike Duplicate Recovery

The vertical bars in the formula above indicate the absolute
value of the difference, hence RPD is always expressed as a
positive value.

The limits for matrix spike compound recovery and RPD are given
in Table 7.  As these limits are only advisory, no further
action by the laboratory is required, however,  frequent
failures to meet the limits for recovery or RPD warrant
investigation by the laboratory, and may result in questions
from the Agency.

                     TABLE 7
            MATRIX SPIKE RECOVERY AND
        RELATIVE PERCENT DIFFERENCE  LIMITS
                  %Recovery   RPD
                  Water	Water
%Recovery  RPD
Soil	Soil
Phenol
2-Chlorophenol
1 , 4 - Dichlorobenzene
N-Nitroso-di-n-propylamine
1 , 2 ,4-Trichlorobenzene
4 - Chloro - 3 -me thylphenol
Acenaphthene
4-Nitrophenol
2 ,4-Dinitrotoluene
Pentachlorophenol
Pyrene
12-110
27-123
36- 97
41-116
39- 98
23- 97
46-118
10- 80
24- 96
9-103
26-127
42
40
28
38
28
42
31
50
38
50
31
26- 90
25-102
28-104
41-126
38-107
26-103
31-137
11-114
28- 89
17-109
35-142
35
50
27
38
23
33
19
50
47
47
36
8.7   Method blank analysis must be performed once for the following, on each
      GC/MS system used to analyze samples,  whichever is most frequent:

      o   Each  Case,  OR

      o   Each  14  calendar day period (7  calendar day period for 14-day data
          turnaround  contracts)  during which samples  in a Case are received
          (said period  beginning with the receipt of  the first sample in that
          Sample Delivery  Group),  OR
      o   Each  20  samples  in a Case,  including  matrix spikes and reanalyses,
          that  are of similar matrix (water  or  soil)  or similar concentration
          (soil only),  OR

      o   Whenever samples are extracted  by  the same  procedure (continuous
          liquid-liquid extraction or sonication).
                                D-59/SV
                                                OLM01.2  1/91

-------
                                                             SECTION IV
      Determine the .concentrations of any target compounds detected in the
     - semivolatile method blank, using the equations in paragraph 8.2.   The
      method blank must contain less than or equal to the Contract Required
      Quantitation Limit (CRQL) of the semivolatile target compounds in
      Exhibit C, except the phthalate esters, which must be less than or equal
      to five times (5x) the CRQL.  For soil/sediment method blanks, CRQL
      value must be adjusted for percent moisture (see Exhibit B).

      If a laboratory method blank exceeds these criteria, the Contractor must
      consider the analytical system to be out of control.  The source of the
      contamination must be investigated and appropriate corrective measures
      MUST be taken and documented before further sample analysis proceeds.
      All samples processed with a method blank that is out of control (i.e.,
      contaminated) MUST be re-extracted and re-analyzed at no additional cost
      to the Agency.  The Laboratory Manager, or his designee, must address
      problems and solutions in the SDG Narrative (Exhibit B).

9.     GC/MS Confirmation of Pesticides and Aroclors

      The requirements for GC/MS confirmation of pesticides and Aroclors are
      given in paragraph 17 of Exhibit D PEST.  When performed, the
      characteristic ions to be used for these analytes are given in Table 8.
      Also see paragraph 3.2 of this section regarding the inclusion of these
      analytes in the semivolatile continuing calibration standard.
                                D-60/SV                              OLM01.0

-------
                                   TABLE 8

                  CHARACTERISTIC IONS FOR PESTICIDES/AROCLORS


Parameter	Primary Ion	Secondary Ion(s)

alpha-BHC                                 183                        181, 109
beta-BHC                                  181                        183, 109
delta-BHC                                 183                        181, 109
gamma-BHC (Lindane)                       183                        181, 109
Heptachlor                                100                        272, 274
Aldrin                                     66                        263, 220
Heptachlor epoxide                        353                        355, 351
Endosulfan I                              195                        339, 341
Dieldrin                                   79                        263, 279
4,4'-DDE                                  246                        248, 176
Endrin                                    263                         82, 81
Endrin ketone                             317                         67, 319
Endrin aldehyde                            67                        250, 345
Endosulfan  II                            337                        339, 341
4,4'-DDD                                  235                        237, 165
Endosulfan sulfate                        272                        387, 422
4,4'-DDT                                  235                        237, 165
Methoxychlor                              227                          228
Chlordane (alpha and/or gamma)            373                        375, 377
Toxaphene                                 159                        231, 233
Aroclor-1016                              222                        260, 292
Aroclor-1221                              190                        222, 260
Aroclor-1232                              190                        222, 260
Aroclor-1242                              222                        256, 292
Aroclor-1248                              292                        362, 326
Aroclor-1254                              292                        362, 326
Aroclor-1260                              360                        362, 394
                                 D-61/SV                              OLM01.0

-------
       EXHIBIT D
  ANALYTICAL METHODS
FOR PESTICIDES/AROCLORS
      D-l/PEST                             OLM01.0

-------
                                  EXHIBIT D


                              Table of Contents

                                                                 Page

SECTION I -   Introduction 	  D-3/PEST

SECTION II -  Sample Preparation and Storage 	  D-6/PEST

              PART A -  Sample Storage and Holding Times 	  D-7/PEST

              PART B -  Sample Preparation for
                        Extractable Pesticides
                        and Aroclors  	  D-8/PEST

SECTION III -  GC/EC Analysis of Pesticides
                 and Aroclors  	  D-35/PEST
                                 D-2/PEST                             OLM01.0

-------
                                  SECTION I

                                 INTRODUCTION
The analytical method that follows is designed to analyze water, sediment and
soil from hazardous waste sites to determine the presence and concentration
of the chlorinated pesticides and Aroclors found in the Target Compound List
(Exhibit C).  The method can be used for determining analyte concentrations
in the range from the contract required quantitation limits (CRQL) to one
million times the CRQL in these matrices.  The method is based on EPA Method
608.

The method  is divided into three sections:  Introduction, Sample Preparation,
and Analysis.  Sample preparation covers sample extraction and cleanup
techniques.  The analysis section contains the specific GC/EC analytical
methods for pesticides and Aroclors.

1.    Summary of the Method

1.1   Continuous liquid-liquid or separatory funnel extraction procedures are
      employed for aqueous samples.  Sonication extraction is required for
      soil/sediment samples (Section II, beginning at 6.2).  The method
      specifies GPC, adsorption column cleanup, and sulfur cleanup techniques
      (Section II, beginning at.7).

1.2   The chlorinated pesticides and Aroclors listed in Exhibit C are
      determined by a two-column GC/EC technique.

1.3   Sample extracts, standards, and blanks must be analyzed within an
      analytical sequence as defined in Section III.  GC/EC analysis begins
      with an initial demonstration of instrument performance and the
      calibration of all pesticides and Aroclors.  Acceptable initial
      calibration is defined in Section III, beginning at 6.  Initial
      calibration must be repeated whenever the calibration verification
      stipulated in Section III, 7, fails or when major instrument
      maintenance or modification is performed.

1.4   An instrument blank and a Performance Evaluation Mixture are analyzed
      no less than once in every 12 hour analytical sequence in order to
      monitor retention times, calibration factors, and column performance.
      Data can be collected only as long as the results for the Performance
      Evaluation Mixtures and instrument blanks fall within the limits
      defined in Section III,  7.  If two consecutive unacceptable Performance
      Evaluation Mixtures are run, all extracts run since the previous
      acceptable Performance Evaluation Mixture must be reanalyzed.
      Additional Performance Evaluation Mixtures and blanks are recommended
      when highly contaminated samples are suspected.

1.5   Calibration and analysis sequence specifications of the GC/EC method
      apply independently to both GC columns.
                                 D-3/PEST                             OLM01.0

-------
1.6   Matrix spike and a matrix spike duplicate analyses must be prepared and
      analyzed at least once for each matrix type or once per Sample Delivery
      Group (SDG), whichever is most frequent.

1.7   Analysis of a sample on both GC columns is required for all samples,
      blanks, matrix spikes, and matrix spike duplicates.

1.8   A single component pesticide is identified if a peak is detected within
      its appropriate retention time window on each of two columns.
      Toxaphene and Aroclors are identified primarily by pattern recognition,
      but RTs of three to five major peaks must also be taken into
      consideration.  Guidance on quantitation of Aroclors is given in
      Section III, paragraph 13.

1.9   Standards for all tentatively identified Aroclors must be run within 72
      hours of the sample analysis in which they were observed.  These
      standards are used to verify identification only; quantitation is based
      on the standard analyzed during initial calibration.

1.10  Quantitative analysis of pesticides/Aroclors must be accomplished by
      the external standard method.  Three-point calibration curves for
      single component analytes and the surrogates must be generated during
      the initial calibration.  A linear response range must be demonstrated
      from the CRQL to a high point at least 16 times greater than the CRQL.
      Single-point calibrations for multicomponent analytes are sufficient
      for quantitation by this method.

1.11  The ECD response for single component analytes must be within the
      three-point calibration range in order for quantitative measurements to
      be made.  The ECD response for the Aroclors/toxaphene must not be
      larger than the response for the high point calibration analysis of the
      single component analytes.  The extracts  must be diluted if the ECD
      response exceeds the calibration range.  Quantitation must be performed
      and reported for both GC columns.

1.12  Absolute retention times (RTs) are used for the identification of
      pesticides/Aroclors.  The absolute retention time window is calculated
      during initial calibration from the mean RT of the standard, using the
      retention time window specifications in Section III, paragraph 8.4.

1.13  The surrogates,  2,4,5,6-Tetrachloro-m-xylene and decachlorobiphenyl,
      must be added to all samples, blanks, matrix spikes, and matrix spike
      duplicates prior to extraction.  The retention time of both surrogates
      must fall within the retention time windows for an analysis to be
      acceptable.   The surrogate recoveries will be determined in all of
      these samples and will be reported to the Agency as a measure of method
      performance.
                                 D-4/PEST                     OLM01.1   12/90

-------
1.14  The criteria in Section III,  paragraph 14,  are used to determine
      whether an analysis is complete or whether  additional cleanup,
      dilution,  or reextraction is  required.

1.15  Resolution difficulties have  been associated with the following pairs
      of compounds using this method:

          o   On a DB-608 or equivalent column,  DDE and Dieldrin; Methoxychlor
             and Endrin ketone; and Endosulfan I  and gamma-Chlordane.

          o   On a DB-1701 or equivalent column, Endosulfan I and gamma-
             Chlordane; and Methoxychlor and Endosulfan sulfate.
                                 D-5/PEST                            OLM01.0

-------
          SECTION II
SAMPLE PREPARATION AND STORAGE
         D-6/PEST                             OLM01.0

-------
                                  SECTION  II


PART A - SAMPLE STORAGE AND HOLDING TIMES.

1.    Procedures for Sample Storage

      The samples must be protected from light and refrigerated at  4°C  (±28C)
      from the time of receipt until 60 days  after delivery of a complete
      reconciled sample data package to the Agency.   After  60  days  the
      samples may be disposed of in a manner  that complies  with all
      applicable regulations.

      The samples must be stored in an atmosphere demonstrated to be  free  of
      all potential contaminants.

      Samples, sample extracts,  and standards must be stored separately.

2.    Procedure for Sample Extract Storage

      Sample extracts must be protected from  light and stored at 4°C  (+2°C)
      until 365 days after delivery of a complete data package to the Agency.

      Sample extracts must be stored in an atmosphere demonstrated  to be  free
      of all potential contaminants.

      Samples, sample extracts,  and standards must be stored separately.

3.    Contract Required Holding Times

      The extraction of water samples by separatory funnel  procedures must be
      completed within five days of the Validated Time of Sample Receipt
      (VTSR).   Extraction of water samples by continuous liquid-liquid
      extraction procedures must be started within five days of VTSR.
      Extraction of soil/sediment samples by  sonication must be completed
      within 10 days of VTSR.

      Analysis of sample extracts must be completed within 40 days  following
      the start of extraction.
                                 D-7/PEST                       OLM01.3   2/91

-------
                                                              SECTION II
PART B - SAMPLE PREPARATION FOR EXTRACTABLE PESTICIDES AND AROCLORS

1.    Summary of Sample Preparation Methods

1.1   Water Samples

      A 1-L volume of sample is spiked with the surrogate solution and is
      extracted with methylene chloride by using a separatory funnel or a
      continuous extractor.  The methylene chloride extract is dried and
      concentrated (5.3).  The extract is then cleaned up by GPC (GPC is
      required when higher molecular weight compounds are present that
      interfere with the analyses of target compounds; GPC is optional for
      all other circumstances), exchanged to hexane, cleaned up by Florisil
      cartridge, and adjusted to a final volume of 1.0 mL or 2.0 mL as
      described beginning at paragraph 7.2.

1.2   Soil/sediment Samples

      A 30 g aliquot of sample is spiked with the surrogate solution and then
      mixed with sodium sulfate and extracted with a 1:1 acetone/methylene
      chloride solvent mixture by sonication.  The extract is then filtered,
      dried, concentrated by K-D, and the solvent exchanged into methylene
      chloride (6.3).  The extract is then cleaned up by GPC (mandatory).
      exchanged to hexane, cleaned up by Florisil cartridge, and adjusted to
      a final volume of 1.0 or 2.0 mL (7.2).

2.    Interferences

2.1   Method interferences may be caused by contaminants in solvents,
      reagents, glassware, and other sample processing hardware.  These
      contaminants lead to discrete artifacts or to elevated baselines in gas
      chromatograms.   Routinely,  all of these materials must be demonstrated
      to be free from interferences under the conditions of the analysis by
      running reagent blanks.   Interferences caused by phthalate esters can
      pose a major problem in pesticide analysis.  Common flexible plastics
      contain varying amounts of phthalates which are easily extracted during
      laboratory operations, so cross-contamination of glassware frequently
      occurs when plastics are handled.  Interferences from phthalates can
      best be minimized by avoiding the use of such plastics in the
      laboratory.

2.2   Matrix interferences may be caused by contaminants that are co-
      extracted from the sample.   The extent of matrix interferences will
      vary considerably from source to source, depending upon the nature of
      the site being sampled.   The cleanup procedures must be used to remove
      such interferences in order to achieve the Contract Required
      Quantitation Limits.
                                 D-8/PEST                      OLM01.2   1/91

-------
                                                              SECTION II


3 .    Apparatus and Materials

      Brand names, suppliers, and part numbers are for illustrative purposes
      only.  No endorsement is implied.  Equivalent performance may be
      achieved using apparatus and materials other than those specified here,
      but demonstration of equivalent performance meeting the requirements of
      this SOW is the responsibility of the Contractor.

3.1   Continuous liquid-liquid extractors - with Teflon or glass connecting
      lines for use with methylene chloride, (Hershberg-Wolf Extractor, Ace
      Glass Company, Vineland, NH, P/N 6841-10 or equivalent).

3.2   Separatory funnel - 2000 mL with Teflon stopcock.

3.3   Apparatus for determining percent moisture

      3.3.1   Oven  - drying.

      3.3.2   Desiccator.

      3.3.3   Crucibles  - porcelain  (optional).

      3.3.4   Aluminum weighing pans  (optional).

3.4   Sonic cell disrupter - Heat Systems,  Ultrasonics, Inc., Model W-385
      (475 watt with pulsing capability, No. 207 3/4-inch tapped disrupter
      horn) or equivalent device with a minimum 375 Watt output capability.
      NOTE:  In order to ensure that sufficient energy is transferred to the
      sample during extraction, the horn must be replaced if the tip begins
      to erode.   Erosion of the tip is evidenced by a rough surface.

3.5   Sonabox (or equivalent) for use with disruptor to decrease noise level.

3.6   Beakers -  400-mL.

3.7   Kuderna-Danish (K-D) apparatus.

      3.7.1   Concentrator  tube -  10-mL, graduated  (Kontes K-570040-1029,  or
              equivalent).

      3.7.2   Evaporative flask -  500-mL (Kontes K-470001-0500, or
              equivalent).

      3.7.3   Snyder column  -  three-ball macro  (Kontes K-503000-0121,  or
              equivalent).

3.8   Funnels and Filter Paper.

      3.8.1   Powder funnels  - 10-cm diameter  (optional), for
              filtration/drying.

      3.8.2   Buchner funnels  - 9-cm diameter,  for  filtration  (optional).



                                 D-9/PEST                       OLM01.2  1/91

-------
                                                              SECTION II


      3.8.3   Filter paper  - No. 41 Whatman (or equivalent), 9-cra circles
              (optional).

3.9   Boiling chips.

      3.9.1   Silicon carbide boiling chips - approximately 10 to 40 mesh.
              Heat  the chips to 400°C for 30 minutes or solvent rinse before
              use.

      3.9.2   Teflon boiling chips (optional) -  solvent rinse the chips
              before use.

3.10  Water bath - heated,  with concentric ring cover,  capable of temperature
      control.   NOTE:   The  water bath should be used in a hood.

3.11  Top loading balance - capable of weighing accurately to  + 0.01  g.

3.12  Balance -  analytical, capable of weighing accurately to  + 0.0001 g.
      The balance must be calibrated with class  S weights once per each 12-
      hour workshift.   The  balances must also be checked annually by  a
      certified technician.

3.13  Nitrogen evaporation  device equipped with a heated bath  that can be
      maintained at 35 to 40°C,  N-Evap by Organomation Associates,  Inc.,
      South Berlin,  MA (or  equivalent).

3.14  Vials and  caps -  2-mL for GC auto sampler.

3.15  Gel permeation chromatography (GPC) cleanup device.   NOTE:  GPC  cleanup
      is required for  all extracts for all soils and for water extracts
      containing higher molecular weight contaminants that interfere  with  the
      analyses  of the  target compounds (see paragraph 7.1.1).

      Gel permeation chromatography system - GPC Autoprep Model 1002  A or  B,
      Analytical Biochemical Laboratories,  Inc., or equivalent.   Systems  that
      perform very satisfactorily also have been assembled from the following
      components - an HPLC  pump,  an auto sampler or a valving  system  with
      sample loops,  and a fraction collector.   All  systems,  whether automated
      or manual, must meet  the  calibration requirements of 7.1.3.

      3.15.1  Chromatographic column - 700 mm x 25 mm i.d. glass column.
              Flow  is upward.  To simplify switching from the UV detector
              during calibration to the GPC collection device during extract
              cleanup, an optional double 3-way valve  (Rheodyne Type 50
              Teflon Rotary Valve #10-262 or equivalent) may be attached so
              that  the column exit flow can be  shunted either to the UV flow-
              through cell or to the GPC collection device.

      3.15.2  Guard column - (Optional) 5 cm, with appropriate fittings to
              connect to the inlet side of the  analytical column (Supelco 5-
              8319 or equivalent).
                                D-10/PEST                      OLM01.2   1/91

-------
                                                              SECTION II
      3.15.3   Bio Beads  (S-X3)  -  200-400 mesh,  70 gm  (Bio-Rad  Laboratories,
               Richmond,  CA,  Catalog  152-2750 or equivalent).   An  additional  5
               gm of Bio  Beads  is  required  if the optional  guard column  is
               employed.   The quality of Bio Beads may vary from lot  to  lot
               because of excessive fines in some lots.   In addition  to  fines
               having a detrimental effect  on chromatography, they also  can
               pass through the column screens  and damage the valve.

      3.15.4   Ultraviolet detector - fixed wavelength (254 run) with  a semi-
               prep flow-through cell.

      3.15.5   Strip chart recorder,  recording  integrator or laboratory  data
               system.

      3.15.6   Syringe -  10-mL  with Luerlok fitting.

      3.15.7   Syringe filter assembly, disposable - Bio-Rad "Prep Disc"
               sample filter assembly #343-0005, 25 mm, and 5 micron  filter
               discs or equivalent.   Check  each batch  for contaminants.   Rinse
               each filter assembly (prior  to use) with methylene  chloride if
               necessary.

      3.15.8   A description of a  manual system assembled from  parts  can be
               found in Wise, R.H., Bishop, D.F., Williams,  R.T. & Austern,
               B.M.  "Gel  Permeation Chromatography in  the GC/MS Analysis  of
               Organics in Sludges" U.S. EPA, Municipal Environmental Research
               Laboratory,  Cincinnati,  Ohio, 45268.

      3.15.9   Glass bottle - 1 liter volume, for use  in  preparation  of  Bio
               beads for  packing into column.

3.16  Florisil  - 500-mg  or 1-g cartridges with stainless steel or Teflon
      frits, (Catalog No.  694-313, Analytichem, 24201 Frampton Ave.,  Harbor
      City, CA, or equivalent).

3.17  Vacuum system for  eluting multiple cleanup cartridges.

      3.17.1   Vac  Elute  Manifold  - Analytichem International,  J.T. Baker, or
               Supelco (or equivalent).  The manifold  design must  ensure that
               there is no contact between  plastics containing  phthalates and
               sample  extracts.

      3.17.2   Vacuum trap made from  a 500-mL sidearm  flask fitted with  a one-
               hole  stopper and glass tubing.

      3.17.3   Vacuum pressure  gauge.

      3.17.4   Rack for holding 10-mL volumetric flasks  in  the  manifold.

3.18  Pyrex glass wool -  rinsed with methylene chloride  and dried before use.

3.19  Bottle or test  tube  - 20-mL with Teflon-lined screw cap for  sulfur
      removal.


                                D-11/PEST                       OLM01.2   1/91

-------
3.20  Drying column - chromatographic column approximately 400 mm long x 19
      mm ID, with coarse frit.  (Substitution of a small pad of disposable
      Pyrex glass wool for the frit will help prevent cross-contamination of
      sample extracts.)

3.21  Glass vials - minimum of 20-mL, with screw cap and Teflon or aluminum
      foil liner.

3.22  Spatula - stainless steel or Teflon.

3.23  pH Paper - wide range, (Hydrion Papers, Microessential Laboratory,
      Brooklyn, NY, or equivalent).

3.24  Pipet - Volumetric 1.00-mL or 2.00-mL (optional).

3.25  Syringe - 1.00-mL or 2.00-mL (optional).

3.26  Flask - Volumetric 10.00-mL.

3.27  Flask - Volumetric 1.00-mL or 2.00-mL (optional).

3.28  Vials - 10-mL, with screw cap and Teflon liner (optional).

3.29  Tube - centrifuge, 12- to 15-mL with 19-mm ground glass joint
      (optional).

3.30  Snyder Column - micro two or three ball with a 19-mm ground glass
      joint.

3.31  Centrifuge - table top (optional).

3.32  Vortex mixer - Genie, Model 550-6, Scientific Industrial, Inc.,
      Bohemia, NY, or equivalent.

3.33  pH Meter with a combination glass electrode.

3.34  Magnetic stirrer motor - Model PC353, Corning Co., Corning, NY, or
      equivalent.

3.35  Magnetic stir bar - Teflon coated, at least 4 cm long.

3.36  Graduated cylinder - 1 L capacity.

4.    Reagents

4.1   Sodium sulfate - granular-anhydrous reagent grade, heated at 400°C for
      4 hours, or at 120°C for 16 hours, cooled in a desiccator, and stored
      in a glass bottle.  Each lot must be extracted with hexane and analyzed
      by GC/EC to demonstrate that it is free of interference before use.
      Baker anhydrous granular, Catalog No. 3375, or equivalent.  CAUTION:
      An open container of sodium sulfate may become contaminated during
      storage in the laboratory.
                                D-12/PEST                             OLM01.0

-------
                                                              SECTION II
4.2   Methylene chloride, hexane, acetone,  toluene,  iso-octane,  and methanol
      (optional) - pesticide quality or equivalent.   It is  recommended that
      each lot of solvent used be analyzed to demonstrate that it is free of
      interference before use.  Methylene chloride must be  certified as acid
      free or must be tested to demonstrate that it is free of hydrochloric
      acid.  Acidic methylene chloride must be passed through basic alumina
      and then demonstrated to be free of hydrochloric acid.

4.3   Mercury - triple distilled, for sulfur clean-up.

4.4   Copper powder (optional) - fine, granular (Mallinckrodt 4649 or
      equivalent).  Copper may be used instead of mercury for sulfur clean-
      up.  Remove oxides by treating with dilute nitric acid, rinse with
      distilled water to remove all traces of acid,  rinse with acetone, and
      dry under a stream of nitrogen.

4.5   Sodium hydroxide solution (10 N) -  Carefully dissolve 40 g of NaOH in
      reagent water and dilute the solution to 100 mL.

4.6   Concentrated sulfuric acid - 18 N.

4.7   Reagent water - defined as a water in which no interferent is observed
      at one-half the CRQL of any pesticide/Aroclor when one liter of the
      reagent water is extracted and prepared by using the same workup
      procedure as for a water sample.

4.8   Ten percent acetone in hexane (v/v) -  prepare by adding 10.0 mL of
      acetone to 90.0 mL of hexane.  NOTE:   Prepare this mixture accurately
      or the results from the Florisil cartridge cleanup will be adversely
      affected.  Water in the acetone also will adversely affect Florisil
      performance.

4.9   Standards

      4.9.1   The Contractor must  provide all standards  to be used with  this
              contract.  These  standards may be used  only after  they have
              been  certified  according  to the procedure  in Exhibit  E.  The
              Contractor must be able to verify that  the standards  are
              certified.  Manufacturer's certificates  of analysis must be
              retained by the Contractor and presented upon  request.

      4.9.2   Stock standard  solutions  (1.00 ug/uL) -  can be prepared  from
              pure  standard materials or purchased  as  certified  solutions.

              4.9.2.1    Prepare stock standard solutions by accurately
                         weighing  about 0.0100 g of pure material.  Dissolve
                         the material in toluene, dilute to volume  in  a  10-mL
                         volumetric flask with toluene or acetone.
                                 D-13/PEST                             OLM01.0

-------
                                                        SECTION II
         4.9.2.2    Transfer the stock standard solutions into a
                   bottle/vial with Teflon-lined cap or septa.  Store
                   at 4°C (±2°C) and protect from light.  Stock
                   standard solutions must be replaced after six months
                   or sooner, if comparison with check standards
                   indicates a problem.

4.9.3    GPC  calibration  solution -  prepare a solution  in methylene
         chloride  that contains  the following analytes in the
         concentrations listed below:

         Analyte                                  mg/mL

         corn oil                                 25
         bis-2-ethylhexyl phthalate               1.0
         methoxychlor                             0.2
         perylene                                 0.02
         sulfur                                   0.08

         NOTE:  Sulfur is not very soluble  in methylene  chloride,
         however,  it is soluble  in warm corn oil.  Therefore,  one
         approach  is to weigh out the corn  oil, warm it  and transfer  the
         weighed amount of sulfur into the  warm corn oil.  Mix it and
         then transfer into a volumetric flask with methylene  chloride,
         along with the other calibration compounds.

         Store the calibration solution in  an amber glass bottle with a
         Teflon lined screw-cap  at 4°C, and protect from light.
         (Refrigeration may cause the corn  oil to precipitate.  Before
         use,  allow the calibration solution to stand at room
         temperature until the corn oil dissolves.)  Replace the
         calibration standard solution every six months, or more
         frequently if necessary.

4.9.4    Surrogate solution -  the surrogates, Tetrachloro-m-xylene and
         Decachlorobiphenyl, are added to all standards, samples, matrix
         spikes, and blanks.  Prepare a surrogate spiking solution of
         0.2  ug/mL of each of the two compounds in acetone.  The
         solution  should  be checked frequently for stability.  The
         solution must be replaced after six months, or  sooner, if
         comparison with  quality control check samples indicates a
         problem.  CAUTION:  Analysts must  allow all spiking solutions
         to equilibrate to room  temperature before use.

4.9.5    Pesticide matrix spiking solution  -  prepare a  spiking solution
         in acetone or methanol  that contains the following pesticides
         in the concentrations specified:
                          D-14/PEST                             OLM01.0

-------
                  Pesticide                ug/mL

                  gamma-BHC (Lindane)       0.5
                  4,4'-DDT                  1.0
                  Endrin                    1.0
                  Heptachlor                0.5
                  Aldrin                    0.5  .
                  Dieldrin                  1.0

              The solution must be prepared every six months, or sooner if
              the solution has degraded or concentrated.

      4.9.6   Florisil cartridge check solution.

              Prepare a solution of  2,4,5-trichlorophenol in acetone, at a
              concentration of 0.1 ug/mL.

      4.9.7   Store all standard solutions in amber glass bottles or vials
              with a teflon-lined screw cap at 4°C (+2°C) and protect from
              light.

5.     Extraction of Water Samples

      Water samples may be extracted by either a separatory funnel procedure
      or a continuous liquid-liquid extraction procedure.   If an emulsion
      prevents acceptable solvent  recovery with the  separatory funnel
      procedure, continuous liquid-liquid extraction must be employed.

5.1   Separatory Funnel Extraction

      5.1.1   Measure out each 1.0 L sample aliquot in a separate graduated
              cylinder.  Measure and record the pH of the sample with wide
              range pH paper and adjust the pH to between 5 and 9 with 10 N
              sodium hydroxide or cone, sulfuric acid, if required.  Samples
              requiring pH adjustment must be noted in the SDG Narrative.
              Place the sample into  a 2-L separatory funnel.

      5.1.2   For each sample selected for matrix spike and matrix spike
              duplicate analyses,  measure out two additional 1-L portions and
              transfer those portions into separate funnels.  Adjust the pH
              of each, if required,  and fortify each with 1.0 mL of matrix
              spike solution before  continuing the extraction.  The frequency
              of MS/MSD analysis is  given in Section III, paragraph 16.

      5.1.3   Using a syringe or a volumetric pipet, add 1.0 mL of the
              surrogate solution to  all water samples, matrix spikes, and
              blanks.

      5.1.4   Add 60 mL methylene chloride to the separatory funnel and
              extract the sample by  shaking the funnel for two minutes, with
              periodic venting to release excess pressure.  Allow the organic
              layer to separate from the water phase for a minimum of 10
              minutes.  If the emulsion interface between layers is more than
              one-third the volume-of the solvent layer, the analyst must

                                D-15/PEST                      OLM01.2   1/91

-------
                                                              SECTION II
              employ-mechanical techniques to complete the phase separation.
              The optimum technique depends upon the sample, and may include
              stirring, filtration of the emulsion through glass wool,
              centrifugation or other physical means.  Drain the methylene
              chloride into a 250 mL Erlenmeyer flask.

      5.1.5   Add a second 60 mL volume of methylene chloride to the sample
              bottle and repeat the extraction procedure a second time,
              combining the extracts in the Erlenmeyer flask.  Perform a
              third extraction in the same manner.

      5.1.6   Prepare a method blank with each group of water samples
              extracted.  For pesticide/Aroclor analyses, a method blank for
              water samples consists of a 1 L volume of reagent water (see
              paragraph 4.7), spiked with the surrogates and carried through
              the entire analytical procedure.  The frequency of method blank
              analysis is given in Section III, paragraph 15.

5.2   Continuous Liquid-Liquid Extraction

      5.2.1   Add methylene chloride (100 to 250 mL) to the bottom of the
              extractor and fill it to a depth of at least one inch above the
              bottom sidearm.

      5.2.2   Measure out each 1.0 L sample aliquot in a separate graduated
              cylinder.  Measure and record the pH of the sample with wide
              range pH paper and adjust the pH to between 5 and 9 with 10 N
              sodium hydroxide or concentrated sulfuric acid, if required.
              Samples requiring pH adjustment must be noted in the SDG
              Narrative.  Place the sample into the continuous extractor.

      5.2.3   With some samples it may be necessary to place a layer of glass
              wool between the methylene chloride and the water layers in the
              extractor to prevent precipitation of suspended solids into the
              methylene chloride during extraction.

      5.2.4   For each sample selected for matrix spike and matrix spike
              duplicate analyses, measure out two additional 1-L portions and
              transfer those portions into separate funnels.  Adjust the pH
              of each, if required, and fortify each with 1.0 mL of matrix
              spike solution before continuing the extraction.  The frequency
              of MS/MSD analysis is given in Section III, paragraph 16.

      5.2.5   Using a syringe or a volumetric pipet, add 1.0 mL of the
              surrogate solution to all water samples, matrix spikes, and
              blanks.

      5.2.6   Adjust the level of methylene chloride in the extractor so that
              the bottom sidearm is half filled with solvent.

      5.2.7   Add sufficient methylene chloride to the distilling flask to
              ensure proper solvent cycling during operation and extract the
              solution for 18 hours.   Allow to cool, then detach the
              distillation flask and label.

                                D-16/PEST                      OLM01.2  1/91

-------
      5.2.8    Prepare  a  method blank with each group of water samples
               extracted.   For pesticide/Aroclor analyses, a method blank for
               water  samples  consists of a 1 L volume of reagent water  (see
               paragraph  4.7), spiked with the surrogates and carried through
               the  entire analytical procedure.  The frequency of method blank
               analysis is given  in Section III, paragraph 15.

5.3   Extract Drying and Concentration

      5.3.1    Assemble a Kuderna-Danish (K-D) concentrator by attaching a 10
               mL concentrator tube to a 500 mL evaporative flask.  Other
               concentration  devices or techniques may be used in place of the
               K-D  if equivalency is demonstrated for all the
               pesticide/Aroclor  target compounds listed in Exhibit C.

      5.3.2    Pour the extract through a drying column containing about 10 cm
               of anhydrous granular sodium sulfate and collect the extract in
               the  K-D  concentrator.  Rinse the Erlenmeyer flask and the
               column with at least two additional 20 to 30 mL portions of
               methylene  chloride to complete the quantitative transfer.

      5.3.3    Add  one  or two clean boiling chips to the evaporative flask and
               attach a three-ball Snyder column.  Pre-wet the Snyder column
               by adding  about 1  mL of methylene chloride to the top of the
               column.  Place the K-D apparatus on a hot water bath (60° -
               80°C)  so that  the  concentrator tube is partially immersed in
               the  hot  water  and  the entire lower rounded surface of the flask
               is bathed  with hot vapor.  Adjust the vertical position  of the
               apparatus  and  the  water temperature as required to complete the
               concentration  in 15 to 30 minutes.  At the proper rate of
               distillation,  the  balls of the column will actively chatter,
               but  the  chambers will not flood with condensed solvent.  When
               the  apparent volume of liquid reaches 3 to 5 mL, remove  the K-D
               apparatus.  Allow  it to drain and cool for at least 10 minutes.
               DO NOT ALLOW THE EVAPORATOR TO GO DRY.

      5.3.4    If no  GPC  cleanup  is required, proceed with the hexane exchange
               described  in paragraph 7.2.  If GPC cleanup is to be used,
               remove the Snyder  column, rinse the flask and its lower  joint
               and  collect the rinsate in the concentrator tube, adjust the
               volume to  10.0 mL  with methylene chloride.  Proceed to 7.1.

6.    Extraction of Soil/Sediment Samples

6.1   Sample Preparation

      6.1.1    Mix  samples  thoroughly, especially composited samples.   Discard
               any  foreign objects such as sticks, leaves and rocks.  Also,
               decant and discard any standing aqueous phase.

      6.1.2    pH Determination - transfer 50 g of soil/sediment to a 100-mL
               beaker.  Add 50 mL of water and stir the solution with a
               magnetic stirrer for 1 hour.  Determine the pH of the sample by
               using  a  glass  electrode and the pH meter while the sample is

                                 D-17/PEST                            OLM01.0

-------
              stirred.  Report pH value on the appropriate data sheet.  If
              the pH of the soil is > 9 or < 5,  document any subsequent
              problems in the analysis related to pH in the SDG Narrative,
              but do not attempt to adjust the pH of the sample.  Discard the
              portion of the sample used for pH determination.

              NOTE:  If insufficient volume of soil is received, use 5 g of
              soil and 5 mL of water for the pH determination and note in the
              SDG Narrative.

      6.1.3   Percent Moisture - Weigh 5 to 10 g of the sediment to the
              nearest 0.01 g into a tared crucible or aluminum weighing pan.
              Determine the weight percent volatilized by drying overnight at
              105°C (hereafter referred to as percent moisture).  After the
              sample is dry, remove the sample and pan and allow them to cool
              in a desiccator before weighing.  Calculate the percent
              moisture according to Equation 1 below.  Concentrations of
              individual analytes will be reported relative to the dry weight
              of sediment.  CAUTION: Gases volatilized from some
              soil/sediment samples may require that this drying procedure be
              carried out in a hood.

              Percent      Wt of Sample - Wt of Dry Sample
              Moisture  -	-	  x 100    EQ. 1
                                  Wt of sample

6.2   Extraction with  Sonication

      6.2.1   Tune the sonicator according to the manufacturer's directions
              prior to extracting samples by this procedure.

      6.2.2   Weigh approximately 30 g of sample (to the nearest 0.1 g) into
              a 250 or 400-mL beaker and add 60 g of anhydrous sodium sulfate
              (granular).

      6.2.3   For a sample to be used for matrix spike and matrix spike
              duplicate analysis, weigh out two additional 30 g (record
              weight to nearest 0.1 g) portions of sample and add 1.0 mL of
              the pesticide matrix spike solution to each soil aliquot.  The
              frequency of MS/MSD analysis is given in Section III, paragraph
              16.

      6.2.4   Add 2.0 mL of surrogate solution to all soil samples, matrix
              spikes,  and blanks by using a volumetric pipet or a syringe.
              Mix the solution well.  The sample and the added sodium sulfate
              should be a homogeneous, granular mixture at this point.
              (Twice as much of the surrogate solution is added to soil
              samples than to water samples because of the increased
              likelihood that the soil sample extracts will require
              dilution).

      6.2.5   Immediately add 80 to 100 mL of 1:1 methylene chloride/acetone
              to the sample.
                                D-18/PEST                     OLM01.1   12/90

-------
      6.2.6    Place  the bottom surface of the sonicator probe about 1/2 inch
               below  the surface of the solvent but above the sediment layer.

      6.2.7    Sonicate for 3 minutes using a 3/4-inch horn at full power
               (output control knob at 10) with pulse on and percent duty
               cycle  knob set at 50 percent.  Do not use a microtip.  NOTE:
               These  settings refer to the Model W-385.  When using a
               sonicator other than Model W-385, refer to the instructions
               provided by the manufacturer for appropriate output settings.

      6.2.8    The  extracted sample can be filtered by using gravity or vacuum
               filtration.

               6.2.8.1    For gravity filtration,  prepare a filtration/drying
                         bed by placing a plug of glass wool in the neck of a
                         10-cm powder funnel and filling the funnel to
                         approximately half its depth (4 or 5 cm) with
                         anhydrous sodium sulfate (80-100 g).   Decant the
                         extract through the packed funnel and collect it in
                         a 500-mL evaporation (K-D) flask.

               6.2.8.2    For vacuum filtration, use Whatman No. 41 paper in
                         the Buchner funnel.  Pre-wet the paper with methy-
                         lene chloride/acetone before decanting the solvent.

      6.2.9    Repeat the extraction two more times with additional 80 to 100
               mL portions of the 1:1 methylene chloride/acetone.  Before each
               extraction, thoroughly mix the solid residue, and make certain
               that the sodium sulfate is free flowing and not a consolidated
               mass.  As required, break up large lumps with a clean spatula.
               Decant and filter the extraction solvent after each sonication
               by using the same funnel described in paragraph 6.2.8.  After
               the  final sonication, pour the entire sample into the funnel
               and  rinse the beaker and funnel with 60 mL of 1:1 methylene
               chloride/acetone.

      6.2.10   Prepare a method blank with each group of soil/sediment samples
               extracted.  For pesticide/Aroclor analyses, a method blank for
               soil/sediment samples consists of 30 g of sodium sulfate  (see
               paragraph 4.1), spiked with the surrogates and carried through
               the  entire analytical procedure.  The frequency of method blank
               analysis is given in Section III, paragraph 15.

6.3   Soil Extract Concentration

      6.3.1    Add  one or two clean boiling chips to the evaporative flask  and
               attach a three-ball macro Snyder column.  Pre-wet the Snyder
               column by adding about 1 mL of methylene chloride to the  top.
               Place  the K-D apparatus on a hot water bath (60 to 80°C)  so
               that the concentrator tube is partially immersed in the hot
               water  and the entire lower rounded surface of the flask is
               bathed with hot vapor.  Adjust the vertical position of the
               apparatus and the water temperature as required to complete  the
               concentration in 15 to 30 minutes.  At the proper rate of

                                D-19/PEST                            OLM01.0

-------
                                                              SECTION II
              distillation, the balls of the column will actively chatter,
              but the chambers will not flood with condensed solvent.  Reduce
              the volume of liquid to less than 10 mL.  Remove the K-D
              apparatus and allow it to drain and cool for at least 10
              minutes.  DO NOT ALLOW THE EVAPORATOR TO GO DRY.

      6.3.2   In order to remove most of the acetone, it is absolutely
              necessary to further reduce the volume of all soil/sediment
              extracts to 1.0 mL.  This is best accomplished using the
              nitrogen evaporation technique (7.3.2).  The presence of
              acetone will cause a dead volume to develop in the GPC column
              and thus will cause loss of surrogates and analytes during GPC
              cleanups.

      6.3.3   Adjust the extract volume to 10.0 mL with methylene chloride.
              Proceed to 7, below, for mandatory GPC and Florisil cartridge
              cleanup of soil extracts.

7.     Extract Cleanup

      There are  three cleanup procedures specified in this  method:  GPC,
      Florisil cartridge,  and sulfur cleanup.   GPC must  be  performed for all
      soil extracts.   GPC must be performed for water extracts that contain
      higher molecular weight contaminants  that interfere with the  analysis
      of the target  analytes.   Florisil cartridge  cleanup is  mandatory for
      all extracts.   Sulfur cleanup  must be performed for all sample extracts
      contaminated with  sulfur.   Blanks and matrix spike and  matrix spike
      duplicate  samples  must be subjected to the same cleanup as  the unspiked
      samples.

7.1   Extract Cleanup by Gel Permeation Chromatography (GPC)

      7.1.1.  GPC cleanup is mandatory for all soil/sediment extracts.  GPC
              must be performed for water extracts that contain higher
              molecular weight contaminants that interfere with the analysis
              of the target analytes.  Gel permeation chromatography (GPC)  is
              a size exclusion cleanup procedure using organic solvents and
              hydrophobic gels in the separation of synthetic macromolecules.
              The packing gel is porous and is characterized by the range or
              uniformity (exclusion range) of that pore size.  In the choice
              of gels, the exclusion range must be larger than the molecular
              size of the molecules to be separated.  A cross-linked divinyl
              benzenestyrene copolymer (SX-3 Bio Beads or equivalent) is
              specified for this method.

              GPC is recommended for the elimination from the sample of
              lipids, polymers, copolymers, proteins, natural resins and
              polymers,  cellular components, viruses, steroids, and dispersed
              high-molecular-weight compounds.  GPC is appropriate for both
              polar and non-polar analytes, therefore, it can be used
              effectively to clean up extracts containing a broad range of
              analytes.
                                D-20/PEST                            OLM01.0

-------
                                                        SECTION II
        Normally, this method is most efficient for removing high
        boiling materials that condense in the injection port area of a
        gas chromatograph (GC) or in the front of the GC column.  This
        residue ultimately will reduce the chromatographic separation
        efficiency or column capacity because of adsorption of the
        target analytes on the active sites.  Pentachlorophenol
        especially is susceptible to this problem.  GPC system
        performance must be validated at least once every seven
        calendar days by demonstrating 80-110 percent recovery of the
        pesticide matrix spike mixture and examining the pattern of
        peaks from an Aroclor 1016/1260 mixture.

7.1.2   GPC Column Preparation

        7.1.2.1    Weigh out 70 gm of Bio Beads (SX-3).   Transfer them
                   to a 1 liter bottle with a Teflon-lined cap or a 500
                   mL separatory funnel with a large bore stopcock,  and
                   add approximately 300 mL of methylene chloride.
                   Swirl the container to ensure the wetting of all
                   beads.  Allow the beads to swell for a minimum of 2
                   hours.  Maintain enough solvent to cover the beads
                   sufficiently at all times.  If a guard column is to
                   be used,  repeat the above with 5 gm of Bio Beads in
                   a 125 mL bottle or a beaker,  using 25 mL of
                   methylene chloride.

        7.1.2.2    Turn the column upside down from its normal
                   position,  and remove the inlet bed support plunger
                   (the inlet plunger is longer than the outlet
                   plunger).   Position and tighten the outlet bed
                   support plunger as near the end as possible, but no
                   closer than 5 cm (measured from the gel packing to
                   the collar).

        7.1.2.3    Raise the end of the outlet tube to keep the solvent
                   in the GPC column, or close the column outlet
                   stopcock.   Place a small amount of solvent in the
                   column to minimize the formation of air bubbles at
                   the base of poured column packing.

        7.1.2.4    Swirl the bead/solvent slurry to get a homogeneous
                   mixture and,  if the wetting was done in a quart
                   bottle,  quickly transfer it to a 500 mL separatory
                   funnel with a large bore stopcock.  Drain the excess
                   methylene chloride directly into the waste beaker,
                   and then start draining the slurry into the column
                   by placing the separatory funnel tip against the
                   column wall.   This will help to minimize bubble
                   formation.   Swirl occasionally to keep the slurry
                   homogeneous.   Drain enough to fill the column.
                   Place the  tubing from the column outlet into a waste
                   beaker below the column, open the stopcock (if
                   attached),  and allow the excess solvent to drain.
                   Raise the  tube to stop the flow, and close the

                          D-21/PEST  '                          OLM01.0

-------
           stopcock when the top of the gel begins to look dry.
           Add additional methylene chloride to just rewet the
           gel.

7.1.2.5    Wipe any remaining beads and solvent from the inner
           walls of the top of the column with a laboratory
           tissue.   Loosen the seal slightly on the other
           plunger assembly (long plunger) and insert it into
           the column.   Make the seal just tight enough so that
           any beads on the glass surface will be pushed
           forward, but loose enough so that the plunger can be
           pushed forward.

           CAUTION:  Do not tighten the seal if beads are
           between the  seal and the glass surface because this
           can damage the seal and cause leakage.

7.1.2.6    Compress the column as much as possible without
           applying excessive force.   Loosen the seal and
           gradually pull out the plunger.  Rinse and wipe off
           the plunger.  Slurry any remaining beads and
           transfer them into the column.  Repeat the step in
           paragraph 7.1.2.5 and reinsert the plunger.   If the
           plunger cannot be inserted and pushed in without
           allowing beads to escape around the seal, continue
           compression  of the beads without tightening the
           seal,  and loosen and remove the plunger as
           described.   Repeat this procedure until the plunger
           is inserted  successfully.

7.1.2.7    Push the plunger until it meets the gel,  then
           compress the column bed about four centimeters.

7.1.2.8    Pack the optional 5 cm column with approximately 5
           gm of preswelled beads (different guard columns may
           require different amounts).  Connect the guard
           column to the inlet of the analytical column.

7.1.2.9    Connect the  column inlet to the solvent reservoir
           (reservoir should be placed higher than the top of
           the column)  and place the column outlet tube in a
           waste container.   Placing a restrictor in the outlet
           tube will force air out of the column more quickly.
           A restrictor can be made from a piece of capillary
           stainless steel tubing of 1/16" OD x 10/1000" ID x
           2".  Pump methylene chloride through the column at a
           rate of 5 mL/min for one hour.

7.1.2.10   After washing the column for at least one hour,
           connect the  column outlet tube, without the
           restrictor,  to the inlet side of the UV detector.
           Connect the  system outlet to the outlet side of the
           UV detector.  A restrictor (same size as the one in
           paragraph 7.1.2.9) in the outlet tube from the UV

                  D-22/PEST                             OLM01.0

-------
                   detector will prevent bubble formation which causes
                   a noisy UV baseline.  The restrictor will not effect
                   flow rate.  After pumping methylene chloride through
                   the column for an additional 1-2 hours, adjust the
                   inlet bed support plunger until approximately 6-10
                   psi backpressure is achieved.  Push the plunger in
                   to increase pressure or slowly pull outward to
                   reduce pressure.

         7.1.2.11   When the GPC column is not to be used for several
                   days, connect the column outlet line to the column
                   inlet to prevent column drying and/or channeling.
                   If channeling occurs, the gel must be removed from
                   the column, reswelled, and repoured as described
                   above.  If drying occurs, methylene chloride should
                   be pumped through the column until the observed
                   column pressure is constant and the column appears
                   wet.  Always recalibrate after column drying has
                   occurred to verify retention volumes have not
                   changed.

                   NOTE:  The description of solvent flow rate and
                   column pressure applies only to the ABC GPC
                   apparatus.  Laboratories using equivalent equipment
                   must develop the parameters for their apparatus
                   which give acceptable performance as described in
                   7.1.4.

         7.1.2.12   The GPC calibration procedure is based on monitoring
                   the elution of standards with a UV detector
                   connected to the GPC column.  Care must be taken to
                   account for any difference in volume (elution time)
                   between the GC column and the detector and between
                   the GPC column and the collection vial.

                   NOTE:  The UV detector calibration procedure
                   described in 7.1.3 is to be used for the analyses of
                   organochlorine pesticides and Aroclors listed in
                   Exhibit C.  IT MUST NOT BE USED FOR THE ANALYSIS OF
                   GC/MS EXTRACTABLES OR OTHER ANALYTES WITHOUT A
                   RECOVERY STUDY.

7.1.3    Calibration of the GPC Column

         7.1.3.1    Using a 10 mL syringe, load sample loop #1 with
                   calibration solution  (paragraph 4.9.3).  With the
                   ABC automated system, the 5 mL sample loop requires
                   a minimum of 8 mL of  the calibration solution.  Use
                   a firm, continuous pressure to push the sample onto
                   the loop.  Switch the valve so that GPC flow is
                   through the UV flow-through cell.
                          D-23/PEST                             OLM01.0

-------
7.1.3.2    Inject the calibration solution  and obtain  a UV
           trace showing a discrete  peak  for each component.
           Adjust the detector and/or  recorder sensitivity  to
           produce a UV trace  that meets  the following
           requirements.   Differences  between manufacturer's
           cell volumes and detector sensitivities may require
           a dilution of the calibration  solution to achieve
           similar results.  An analytical  flow-through
           detector cell will  require  a much less concentrated
           solution than the semi-prep cell and, therefore,  the
           analytical cell is  not acceptable for use.

           o Peaks must be observed and should be symmetrical
             for all compounds in the calibration solution.

           o Corn oil and phthalate peaks must exhibit >8S%
             resolution.

           o Phthalate and methoxychlor peaks must exhibit >85%
             resolution.

           o Methoxychlor and perylene peaks must exhibit >85%
             resolution.

           o Perylene and sulfur peaks must not be saturated
             and must exhibit >90% baseline resolution.

7.1.3.3    Determine the elution times for  the phthalate,
           methoxychlor,  and perylene. Phthalate will elute
           first,  perylene,  last.

7.1.3.4    Choose a "DUMP" time which  removes > 85 percent  of
           the  phthalate.   Choose a  "COLLECT" time so  that  > 95
           percent of the methoxychlor is collected, and
           continue to collect until just prior to the elution
           time of sulfur.   Use a "WASH"  time of 10 minutes.

7.1.3.5    NOTE:   The DUMP and COLLECT times must be adjusted
           to compensate for the difference in volume  of  the
           lines between the detector  and the collection  flask.

7.1.3.6    Verify the flow rate by collecting column eluate for
           10 minutes in a graduated cylinder and measure the
           volume,  which should be 45-55  mL (4.5-5.5 mL/min).
           If the flow rate is outside of this range,
           corrective action must be taken  to achieve  this  flow
           rate.   Once the flow rate is within the range  of
           4.5-5.5 mL/min,  record the  column pressure  (should
           be 6-10 psi)  and room temperature.  Changes in
           pressure,  solvent flow rate, and temperature
           conditions can affect analyte  retention times  and
           must be monitored.   If the  flow  rate and/or column
           pressure do not fall within the  above  ranges,  a new
           column should be prepared.   A  UV trace  that does not

                  D-24/PEST                      OLM01.2   1/91

-------
                                                        SECTION II
                   meet the criteria in paragraph 7.1.3.2 would also
                   indicate that a new column should be prepared.   It
                   may be necessary to obtain a new lot of Bio Beads if
                   the column fails all the criteria.

         7.1.3.7    Reinject the calibration solution after appropriate
                   collect and dump cycles have been set, and the
                   solvent flow and column pressure have been
                   established.

                   7.1.3.7.1    Measure and record the volume of
                                colleqted GPC eluate in a graduated
                                cylinder.   The volume of GPC eluate
                                collected for each sample extract
                                processed may be used to indicate
                                problems with the system during sample
                                processing.

                   7.1.3.7.2    The retention times for bis(2-
                                ethylhexyl)phthalate and perylene must
                                not vary more than ±5% between
                                calibrations.  If the retention time
                                shift is >5%, take corrective action.
                                Excessive retention time shifts are
                                caused by the following:

                               o  Poor  laboratory  temperature  control  or
                                  system leaks.

                               o  An  unstabilized  column that  requires
                                  pumping  methylene  chloride  through it
                                  for several more hours or  overnight.

                               o  Excessive laboratory temperatures
                                  causing  outgassing of the  methylene
                                  chloride.

         7.1.3.8    Analyze a GPC blank by loading 5 mL of methylene
                   chloride into the GPC.  Concentrate the methylene
                   chloride that passes through the system during the
                   collect cycle using a Kuderna-Danish  (KD)
                   evaporator.  Exchange the solvent to hexane and
                   analyze the concentrate by GC/EC.  If the blank
                   exceeds one half the CRQL of any analyte, assuming
                   that the blank represents the extract from a one
                   liter water sample, pump additional methylene
                   chloride through the system for 1-2 hours.  Analyze
                   another GPC blank to ensure the system is
                   sufficiently clean.  Repeat the methylene chloride
                   pumping if necessary.

7.1.4    GPC Calibration Check

         No Florisil cleanup is used  in the GPC  calibration  check.

                          D-25/PEST                       OLM01.2   1/91

-------
7.1.4.1 ...  At least once every 7  days,  the  calibration of the
           GPC must be  verified with  two  check mixtures.   The
           first mixture is  prepared  by concentrating 2.0 mL of
           the matrix spiking solution  (paragraph 4.9.5)  to
           less than 1  mL under a stream  of nitrogen (7.3.2),
           and adjusting the final volume to 10.0 mL with
           methylene chloride.  The second  mixture is prepared
           with 2 ug of Aroclor 1016  and  2  ug of Aroclor  1260
           in a final volume of 10.0  mL methylene chloride.

7.1.4.2    Load the first 5.0 mL  sample loop by using a 10 mL
           syringe containing 8 mL of the diluted pesticide
           matrix spike solution  (paragraph 7.1.4.1).   The
           Aroclor mixture is loaded  into Loop 2 in the same
           manner.  Fractions are collected in an auto sequence
           by using the GPC  program established by the UV
           detector calibration procedure (Paragraph 7.1.3).

7.1.4.3    The collected GPC calibration  fraction is
           transferred  to a  K-D apparatus,  and the collection
           vessel is rinsed  with  two  additional 10-mL portions
           of methylene chloride  to complete the transfer.  The
           volume of methylene chloride is  reduced (described
           in 7.3.2).   After cooling, the solvent is exchanged
           to hexane according to the instruction in 7.2. The
           final volume is adjusted to  10.0 mL,  and the sample
           is analyzed  by GC according  to the procedures  in
           Section III.   The analysis must  be performed on at
           least one of the  GC columns  used for samples
           analysis.

7.1.4.4    The pattern  of the Aroclor quantitation peaks  and
           the recovery of each single  component analyte  must
           be determined for evaluation and reporting purposes.
           If the recovery of each of the single component
           analytes is  80 to 110  percent  and if the Aroclor
           pattern is the same as with  previously run
           standards, then the analyst  may  continue to use the
           column.  If  recoveries are out of the acceptance
           window or if changes in the  relative peak heights of
           the patterns of the Aroclor  are  observed, the  column
           must be replaced  and recalibrated according to the
           instructions in 7.1.3.

7.1.4.5    Some samples may  contaminate the SX-3 Bio Beads and
           change the retention volume  of the GPC column.
           Therefore system  calibration and analyte recovery
           must be checked whenever a sample causes significant
           discoloration of  the GPC column.  Even if no
           darkening is visible,  GPC  calibration must be
           checked not  less  than  once every seven days.  In
           many cases,  the SX-3 Bio Beads may be used for
           several months as long as  the  column calibration  and
           flow rate remain  constant.

                  D-26/PEST                     OLM01.1   12/90

-------
7.1.5   Daily UV calibration check (optional)

        The calibration of the GPC may be monitored daily by use of the
        UV-GPC calibration solution (paragraph 4.9.3) and the UV
        Detector Calibration Procedure (7.1.3).  The UV detector should
        be used to monitor the elution times for the phthalate,
        methoxychlor, and perylene, in that.order.  The precalibrated
        GPC program should "DUMP" > 85 percent of the phthalate and
        should "COLLECT" > 95 percent of the methoxychlor and perylene.
        Significant changes in elution times of the analytes (e.g., >
        0.5 minutes) indicate that the column is out of calibration and
        must be recalibrated or replaced.

7.1.6   Sample Extract Cleanup

        It is very important to have consistent laboratory temperatures
        during an entire GPC run, which could be 24 hours or more.  If
        temperatures are not consistent,  retention times will shift,
        and the dump and collect times determined by the calibration
        standard no longer will be appropriate.  The ideal laboratory
        temperature to prevent outgassing of the methylene chloride is
        22°C.

        7.1.6.1    In order to prevent overloading of the GPC column,
                   highly viscous sample  extracts must be diluted prior
                   to cleanup.   Any sample extract with a viscosity
                   greater than that of a 1:1 glycerol:water solution
                   must be diluted and loaded into several loops.
                   Similarly,  extracts containing more than 500 mg of
                   nonvolatile residue per 5 mL of extract must be
                   diluted and loaded into several loops.   The
                   nonvolatile residue may be determined by evaporating
                   a 100 uL aliquot of the extract to dryness in a
                   tared aluminum weighing pan,  or other suitable
                   container.

        7.1.6.2    Particles greater than 5 micron may scratch the
                   valve,  which may result in a system leak and cross
                   contamination of sample extracts in the sample
                   loops.   To avoid such  problems,  filter the extract
                   through a 5 micron filter disc by attaching a
                   syringe filter assembly containing the filter disc
                   to a 10 mL syringe. Draw the sample extract through
                   the filter assembly and into the 10 mL syringe.
                   Disconnect the filter  assembly before transferring
                   the sample extract into a small glass container,
                   e.g.,  a 15 mL culture  tube with a Teflon lined screw
                   cap.   Alternatively, draw the extract into the
                   syringe without the filter assembly.   Attach the
                   filter assembly and force the extract through the
                          D-27/PEST                      OLM01.2  1/91

-------
           filter and into the glass container.   Draw a minimum
           of 8 mL of extract into a 10 mL syringe.
           INTRODUCTION OF PARTICIPATES OR GLASS WOOL INTO THE
           GPC SWITCHING VALVES MAY REQUIRE FACTORY REPAIR OF
           THE APPARATUS.

7.1.6.3    Prior to loading samples,, put the GPC into the
           "LOAD" mode, set the instrument terminal for the
           number of loops to be loaded, and set the "DUMP",
           "COLLECT", and "WASH" times for the values
           determined by the calibration procedure described in
           7.1.3.

7.1.6.4    Attach the syringe to the turn lock on the injection
           port.  Use firm,  continuous pressure to push the
           sample onto the 5-mL sample loop.  If the sample is
           difficult to load, some part of the system may be
           blocked.  Take appropriate corrective action.  If
           the back pressure is normal (6-10 psi) the blockage
           is probably in the valve.  Blockage may be flushed
           out of the valve by reversing the inlet and outlet
           tubes and pumping solvent through the tubes (this
           should be done before sample loading).

           NOTE:  Approximately 2 mL of the extract remains in
           the lines between the injection port and the sample
           loop; excess sample also passes through the sample
           loop to waste.

7.1.6.5    After loading a loop, and before removing the
           syringe from the injection port, index the GPC to
           the next loop.  This will prevent loss of sample
           caused by unequal pressure in the loops.

7.1.6.6    After loading each sample loop, wash the loading
           port with methylene chloride in a PTFE wash bottle
           to minimize cross contamination.  Inject
           approximately 10 mL of methylene chloride to rinse
           the common tubes.

7.1.6.7    After loading all the sample loops, index the GPC to
           the 00 position,  switch to the "RUN" mode and start
           the automated sequence.  Process each sample using
           the collect and dump cycle times established in
           7.1.3.

7.1.6.8    Collect each sample in a 250-mL Erlenmeyer flask,
           covered with aluminum foil to reduce solvent
           evaporation, or directly into a Kuderna-Danish
                  D-28/PEST                            OLM01.0

-------
                         evaporator.  Monitor sample volumes collected.
                         Changes in sample volumes collected may indicate one
                         or more of the following problems:

                         o  Change  in  solvent flow rate, caused by channeling
                            in the  column  or changes in column pressure.

                         o  Increase in column operating pressure due  to  the
                            absorption of  particles or gel fines onto  either
                            the guard  column or the analytical column  gel,  if
                            a  guard column is not used.

                         o  Leaks in "the system or significant variances  in
                            room temperature.

               7.1.6.9    After the appropriate GPC fraction has been
                         collected for each sample,  the solvent must be
                         exchanged to hexane as described in 7.2.

               7.1.6.10   Any samples that were loaded into two or more loops
                         must be recombined before proceeding to 7.2.

7.2   Solvent exchange into hexane

      This procedure applies to both extracts  of  water samples and extracts
      of soil samples.

      7.2.1    With  the extract  in  a k-D apparatus, remove the Snyder  column,
               add 50 mL of hexane  and  a new boiling chip, and reattach the
               Snyder column.   Pre-wet  the  column by adding about 1 mL of
               hexane to the top.   Concentrate the solvent extract as
               described previously.  When  the apparent volume of liquid
               reaches 3 to 5 mL, remove the K-D apparatus and allow it to
               drain and cool for at least  10 minutes.  DO NOT ALLOW THE
               EVAPORATOR TO GO  DRY.

      7.2.2    Remove the Snyder column; using 1 to 2 mL of hexane, rinse the
               flask and its lower  joint into the concentrator tube.   Complete
               quantitative transfer of the extract to a 10 mL vial by using
               hexane.

      7.2.3    For samples which have not been subjected to GPC cleanup,
               adjust the volume of the hexane extract to 10.0 mL.  For
               samples which have been  subjected to GPC cleanup, concentrate
               the hexane extract to 5.0 mL using nitrogen evaporation, as
               described in  7.3.2.   Proceed to 7.3 for Florisil cartridge
               cleanup.

7.3   Florisil Cartridge Procedure

      Florisil cartridge cleanup is required for all extracts.  Cleanup
      significantly reduces matrix interferences caused by polar compounds.
                                 D-29/PEST                       OLM01.3  2/91

-------
7.3.1   Cartridge Performance Check -  every lot number of Florisil
        cartridges must be tested by the following procedure before
        they are used for sample cleanup.  Add 0.5 mL of 2,4,5-
        trichlorophenol solution (0.1 ug/mL in acetone) and 0.5 mL of
        Standard Mixture A, midpoint concentration (Section III,
        paragraph 3.3) to 4 mL of hexane.  Reduce the final volume to
        0.5 mL using nitrogen (paragraph 7.3.2).  Place the mixture
        onto the top of a washed Florisil cartridge (paragraph
        7.3.4.4), and elute it with 9 mL of hexane/acetone
        [(90:10)(V/V)].  Use two additional 1-mL hexane rinses to
        ensure quantitative transfer of standard from the cartridge.
        Reduce the final volume to 1.0 mL using nitrogen (7.3.2) and
        analyze the solution by GC/EC using at least one of the GC
        columns specified for sample analyses.  The recovery of each
        analyte must be determined for evaluation and reporting
        purposes.  Calculate the percent recovery using the equation
        below.  The lot of Florisil cartridges is acceptable if all
        pesticides are recovered at 80 to 120 percent, if the recovery
        of trichlorophenol is less than 5%, and if no peaks interfering
        with the target analytes are detected.

                                          Qd                      EQ.  l

                      Percent  Recovery =  	  x  100

                                          Qa

         Where,

         Q
-------
                                               SECTION  II
7.3.3.2    Place one Florisil cartridge  into  the  vacuum
           manifold for each sample  extract.

7.3.3.3    The required Florisil cartridge size and the final
           volume of the extract after Florisil cleanup are a
           function of the GC autosampler that a  laboratory
           uses.  If the autosampler operates reliably with 1.0
           mL of sample extract, then a  500-mg cartridge is
           used and the required final volume is  1.0 mL.  If
           the autosampler requires  more sample,  prepare 2.0 mL
           of sample extract using a 1-g cartridge.   Manual
           injection requires only a 1.0 mL final extract
           volume and a 500-mg cartridge.

7.3.3.4    Prior to cleanup of samples,  the cartridges must be
           washed with hexane/acetone (90:10).  This is
           accomplished by placing the cartridge  in the vacuum
           manifold, by pulling a vacuum, and by  passing at
           least 5 mL of the hexane/acetone solution through
           the cartridge.   While the cartridges are being
           washed, adjust the vacuum applied to each cartridge
           so that the flow rate through each cartridge is
           approximately equal.   DO  NOT  ALLOW THE CARTRIDGES TO
           GO DRY AFTER THEY HAVE BEEN WASHED.

7.3.3.5    After the cartridges in the manifold are washed,  the
           vacuum is released,  and a rack containing labeled
           10-mL volumetric flasks is placed inside the
           manifold.  Care must be taken to ensure that the
           solvent line from each cartridge is placed inside of
           the appropriate volumetric flask as the manifold top
           is replaced.

7.3.3.6    After the volumetric flasks are in place, vacuum to
           the manifold is restored, and a volume of extract
           equal to the required final volume (1.0 or 2.0 mL)
           from each sample, blank or matrix spike extract is
           transferred to the top frit of the appropriate
           Florisil cartridge.

7.3.3.7    Because the volumes marked on concentrator tubes are
           not necessarily accurate  at the 1-mL level, the use
           of a syringe or a volumetric  pipet is  required to
           transfer the extract to the cleanup cartridge.

7.3.3.8    The pesticides/Aroclors in the extract concentrates
           are then eluted through the column with 8 mL of
           hexane/acetone (90:10) and are collected into the
           10-mL volumetric flasks held  in the rack inside the
           vacuum manifold.
                  D-31/PEST                      OLM01.2   1/91

-------
               7.3.3.9    Transfer the eluate in each volumetric flask to a
                         clean centrifuge tube or 10-mL vial.  Use two
                         additional 1-mL hexane rinses to ensure quantitative
                         transfer of the cartridge eluate.

               7.3.3.10   Concentrate the extract to 1.0 or 2.0 mL as required
                         in paragraph 7.3.4.3 by using either nitrogen
                         blowdown (7.3.2) or a micro-Snyder column.  Measure
                         the final volume with a syringe or by transferring
                         the extract to a volumetric flask.

               7.3.3.11   Sulfur contamination will cause a rise in the
                         baseline of the chroraatogram that may interfere with
                         the analyses of the ,later eluting pesticides.  If
                         crystals of sulfur are evident or if the presence of
                         sulfur is suspected, proceed to 7.4.  Sample
                         analyses showing the presence of sulfur are not
                         acceptable and must be cleaned up and reanalyzed.

               7.3.3.12   If sulfur is not present, transfer the sample to a
                         GC vial and label the vial.  The extract is ready
                         for GC/EC analysis.  Proceed to Section III.  Store
                         the extracts at 4°C (+2°C)in the dark.

7.4   Sulfur Removal

      Sulfur can be removed by one of two methods, according to laboratory
      preference.  Interference which is due to sulfur is not acceptable.   If
      the sulfur concentration is such that crystallization occurs in the
      concentrated extract, centrifuge the extract to settle the sample
      extract with a disposable pipette, leaving the excess sulfur in the
      centrifuge tube.   Transfer the extract to a clean centrifuge tube or
      clean concentrator tube before proceeding with further sulfur cleanup.

      7.4.1    If only part of a set of  samples requires sulfur cleanup, then
               two method blanks are required for that set:  one that  is
               shaken with mercury or copper, and one that  is not.

               Sulfur cleanup blank -  add 1.0 mL of surrogate  to  10 mL of
               hexane in a clean centrifuge tube or 10-mL vial.  Concentrate
               the solution to 2.0 mL by using either nitrogen  evaporation or
               a  micro Snyder column.  The concentrated volume  of  the blank
               must be the same as the final volume of the  samples associated
             .  with the blank.  Measure  the volume with a syringe or by
               transferring the solution to a volumetric flask.  Proceed with
               the sulfur removal using  the same technique  (mercury or copper)
               as the samples associated with the blank.
                                D-32/PEST                             OLM01.0

-------
                                                        SECTION II
7.4.2   Mercury  technique

        Add one  to  three drops of mercury to each hexane extract in a
        clean vial.  Tighten the top on the vial and agitate the sample
        for 30 seconds.  Filter or centrifuge the extract.  Pipet the
        extract  to  another vial and leave all solid precipitate and
        liquid mercury.  If the mercury appears shiny, proceed to
        Section  III and analyze the extract.  If the mercury turns
        black, repeat sulfur removal as necessary.  CAUTION:  Waste
        containing  mercury should be segregated and disposed of
        properly.

        NOTE:  Mercury is a highly toxic metal and therefore must be
        used with great care.  Prior to using mercury, it is
        recommended that the analyst become acquainted with proper
        handling and cleanup techniques associated with this metal.

7.4.3   Copper technique

        Add approximately 2 g of cleaned copper powder to the extract
        in the centrifuge or concentrator tube.  (2 g will fill the
        tube to about the 0.5 mL mark).  Mix the copper and extract for
        at least 1 minute on a mechanical shaker.  Separate the extract
        from the copper powder by drawing off the extract with a
        disposable pipet, and transfer the extract to a clean vial.
        The extract transferred to the vial still represents the 2.0 mL
        final volume.  The separation of the extract from the copper
        powder is necessary to prevent degradation of the pesticides.
        If the copper appears bright, proceed to Section III and
        analyze the extract.  If the copper changes color, repeat the
        sulfur removal procedure as necessary.
                          D-33/PEST                            OLM01.0

-------
               SECTION III
GC/EC ANALYSIS OF PESTICIDES AND AROCLORS
              D-34/PEST                            OLM01.0

-------
                                 SECTION III

                  GC/EC ANALYSIS OF PESTICIDES AND AROCLORS
1.     Summary of GC/EC Analysis

1.1   The analysis of samples is accomplished by using a wide-bore (0.53 mm
      ID) fused silica capillary column.

1.2   Sample extracts, standards,  and blanks must be analyzed within an
      analytical sequence as defined in 5.   GC/EC analysis begins with an
      initial demonstration of instrument performance and the calibration of
      all pesticides and Aroclors.   Acceptable initial calibration is defined
      in paragraph 6.  Initial calibration must be repeated whenever the
      calibration verification stipulated in paragraph 7 fails,  or when major
      instrument maintenance or modification is performed.

1.3   An instrument blank, a Performance  Evaluation Mixture,  and a second
      instrument blank and the midpoint concentration of Individual Standard
      Mixtures A and B are analyzed no less than once in every 12 hour
      analytical sequence in order to monitor retention times,  calibration
      factors, and column performance.  Data can be collected only as long as
      the results for these standards and instrument blanks fall within the
      limits defined in paragraph 7.  If two consecutive unacceptable
      standards are run, all extracts run since the previous acceptable
      standard must be reanalyzed.   Additional standards and blanks are
      recommended when highly contaminated samples are suspected.

1.4   Calibration and run sequence specifications of the GC/EC method apply
      independently to both GC columns.

1.5   Matrix spike and a matrix spike duplicate analyses must be prepared and
      analyzed at least once for each matrix type or once per Sample Delivery
      Group (SDG), whichever is most frequent.

1.6   Analysis of a sample on both GC columns is required for all samples,
      blanks, matrix spikes, and matrix spike duplicates.

1.7   A single component pesticide is identified if a peak is detected within
      its appropriate retention time window on each of two columns.
      Toxaphene and Aroclors are identified primarily by pattern recognition,
      but RTs of three to five major peaks must also be taken into
      consideration.  Guidance on quantitation of Aroclors is given beginning
      at 13.9.

1.8   Standards for all tentatively identified Aroclors must be run within 72
      hours of the sample analysis in which they were observed.  These
      standards are used to verify identification only; quantitation is based
      on the standard analyzed during initial calibration.

1.9   Quantitative analysis of pesticides/Aroclors must be accomplished by
      the external standard method.  Three-point calibration curves for
      single component analytes and the surrogates must be generated during

                                D-35/PEST                             OLM01.0

-------
      the initial calibration.  A linear response range must be demonstrated
      from the CRQL to a high point at least 16 times greater than the CRQL.
      Single-point calibrations for multicomponent analytes are sufficient
      for quantitation by this method.

1.10  The ECD response for single component analytes must be within the
      three-point calibration range in order for quantitative measurements to
      be made.  The ECD response for the Aroclors/toxaphene must not be
      larger than the response for the high point calibration analysis of the
      single component analytes.  The extracts must be diluted if the ECD
      response exceeds the calibration range.  Quantitation must be performed
      and reported for both GC columns.

1.11  Absolute retention times (RTs) are used for the identification of
      pesticides/Aroclors.  The absolute retention time window is calculated
      during initial calibration from the RT of the standard, using the
      retention time window specifications in Paragraph 8.4.

2.    Gas Chromatograph/Election Capture Detector (GC/EC)

2.1   Gas Chromatograph

      2.1.1   The gas chromatograph  (GC)  system must adequately regulate
              temperature in  order  to give a reproducible  temperature program
              and have a  flow controller  that maintains a  constant column
              flow rate throughout  temperature program operations.  The
              system must be  suitable for splitless injection and have all
              required accessories  including syringes, analytical columns,
              and gases.

      2.1.2   Gas chromatographs that are available from some manufacturers
              may have difficulty in meeting certain method QC requirements
              because of  Endrin and DDT breakdown in the injector.  This
              problem can be  minimized by operating the injector at 200-
              205°C, using a  Pvrex  (not quartz) methyl silicone deactivated
              injector liner,  and deactivating any metal parts in the
              injector with dichlorodimethyl silane.  In some cases, using  a
              0.25 inch packed column injector converted for use with 0.53  mm
              capillary columns works better than a Grob-type injector.   If a
              Grob-type injector is used, a 4 mm liner may be required to
              meet breakdown  criteria.

2.2   Gas Chromatograph Columns

      2.2.1   Two wide-bore (0.53 mm  ID)  fused silica GC columns are
              required.  A separate detector is required for each column.
              The specified analytical columns are a DB-1701, 30 m x 0.53 mm
              ID, 1.0 urn  film thickness,  (J&W Scientific,  Folsom, CA, or
              equivalent), and a DB-608,  30 m x 0.53 mm ID, 0.5 to 1.0 urn
              film thickness  (J&W Scientific, or equivalent).  Equivalent
              columns may be  employed if  they meet the requirements for
              resolution, initial calibration, and calibration verification
              listed in this  section.
                                D-36/PEST                      OLM01.1  12/90

-------
                                                             SECTION III
      2.2.2-    Columns are mounted in 0.25-inch injector ports by using glass
               adapters available from a variety of commercial sources (J&W
               Scientific, Supelco, Inc., Hewlett-Packard, Varian, Inc.,
               Perkin Elmer, or equivalent).  The two columns may be mounted
               into a single injection port with a tee adapter (Supelco, Inc
               Bellefonte, PA, Catalog No. 2-3660, or equivalent).  Use of
               this adapter allows simultaneous injection onto both columns.
               The laboratory should follow manufacturer's recommendations for
               mounting 0.53 mm capillary columns in injector ports.

2.3   The carrier gas for routine applications  is helium.   Laboratories may
      choose to use hydrogen as a carrier gas,  but they must clearly identify
      its use in the SDG Narrative and on all divider pages preceding raw
      chromatographic data in submissions to the  Agency.   Laboratories that
      choose to use hydrogen are advised to exercise caution in its use.   Use
      of a hydrogen leak detector is highly recommended when hydrogen is used
      as the carrier gas.  All GC carrier gas lines must be constructed from
      stainless steel or copper tubing.   Non-polytetrafluoroethylene (PTFE)
      thread sealants or flow controllers with  rubber components are not to
      be used.

2.4   Electron Capture Detector - the make-up gas must be P-5,  P-10
      (argon/methane) or nitrogen according to  the instrument specification.
      The linearity of the response of the  ECD  may be greatly dependent on
      the flow rate of the make-up gas.   Care must be taken to maintain
      stable and appropriate flow of make-up gas  to the detector.   The GC/EC
      system must be in a room in which the atmosphere has been demonstrated
      to be free of all contaminants which  may  interfere with the analysis.
      The instrument must be vented to outside  the facility or to a trapping
      system which prevents the release of contaminants into the instrument
      room.

2.5   Data System -  a data system must be interfaced to the GC/EC.   The data
      system must allow the continuous acquisition of data throughout the
      duration of the chromatographic program and must permit,  at the
      minimum, the output of time vs.  intensity (peak height or peak area)
      data.  Also,  the data system must be able to rescale chromatographic
      data in order to report chromatograms meeting the requirements listed
      within this method.

3.     Calibration Standards

3.1   Resolution Check Mixture - prepare the mixture of pesticides in hexane
      or iso-octane at the concentrations listed below.  The mixture must be
      prepared every six months, or sooner if the solution has degraded or
      concentrated.

       gamma-Chlordane       10.0 ng/mL   Endrin ketone          20.0 ng/mL
       Endosulfan I          10.0 ng/mL   Methoxychlor          100.0 ng/mL
       p,p'-DDE              20.0 ng/mL   Tetrachloro-m-xylene   20.0 ng/mL
       Dieldrin              20.0 ng/mL   Decachlorobiphenyl     20.0 ng/mL
       Endosulfan sulfate    20.0 ng/mL
                                D-37/PEST                      OLM01.1   12/90

-------
3.2
3.3
Performance Evaluation Mixture (PEM) - prepare the PEM in hexane or
iso-octane at the concentration levels listed below.   The PEM must be
prepared weekly, or more often if the solution has degraded or
concentrated.
       gamma-BHC
       alpha-BHC
       4,4'-DDT
       beta-BHC
                       10.0 ng/mL
                       10.0 ng/mL
                      100.0 ng/mL
                       10.0 ng/mL
      Endrin
      Methoxychlor
      Tetrachloro-m-xylene
      Decachlorobiphenyl
 50.0 ng/mL
250.0 ng/mL
 20.0 ng/mL
 20.0 ng/mL
Individual Standard Mixtures A and B - the single component pesticide
standards must be prepared in hexane or iso-octane at three
concentrations for each analyte,  including the surrogates.   Two
separate calibration mixtures, A and B, (listed below) are used to
ensure that each peak is adequately resolved.   The low point
concentration corresponds to the CRQL for each analyte.  The midpoint
concentration must be 4 times the low point concentration.   The high
point concentration must be at least 16 times  that of the low point,
but a higher concentration may be chosen by the Contractor.  The high
point concentration defines the upper end of the concentration range
for which the concentration is valid.  The solution must be prepared
every six months, or sooner if the solution has degraded or
concentrated.

Individual Standard Mixture A - Low Point Concentration
            alpha-BHC
            Heptachlor
            gamma-BHC
            Endosulfan  I
            Dieldrin
            Endrin
            p,p'-DDD
            p,p'-DDT
            Methoxychlor
            Tetrachloro-m-xylene
            Decachlorobiphenyl
                               5.0 ng/mL
                               5.0 ng/mL
                               5.0 ng/mL
                               5.0 ng/mL
                              10.0 ng/mL
                              10.0 ng/mL
                              10.0 ng/mL
                              10.0 ng/mL
                              50.0 ng/mL
                               5.0 ng/mL
                              10.0 ng/mL
      Individual  Standard Mixture  B-  Low  Point  Concentration
           beta-BHC
           delta-BHC
           Aldrin
           Heptachlor  expoxide
           aIpha-Ch1ordane
           gamma-Chlordane
           p,p'-DDE
           Endosulfan  sulfate
           Endrin aldehyde
           Endrin ketone
           Endosulfan  II
           Tetrachloro-m-xylene
           Decachlorobiphenyl
                               5,
                               5
                               5
                               5,
                               5,
                               5,
  .0 ng/mL
  .0 ng/mL
  .0 ng/mL
  .0 ng/mL
  .0 ng/mL
  .0 ng/mL
10.0 ng/mL
10.0 ng/mL
10.0 ng/mL
10.0 ng/mL
10.0 ng/mL
 5.0 ng/mL
10.0 ng/mL
                                D-38/PEST
                                                        OLM01.1   12/90

-------
 3.4   Multicomponent Standards - Toxaphene and Aroclor standards must be
      prepared, individually except for Aroclor 1260 and Aroclor 1016 which
      may be combined in one standard mixture.  The calibration standards for
      the Aroclors must be prepared at concentrations of 100 ng/mL, except
      for Aroclor 1221 which must be prepared at 200 ng/mL.  Toxaphene must
      be prepared at 500 ng/mL.  All multicomponent standards must contain
      the surrogates at 20.0 ng/mL.  The Aroclor and Toxaphene solutions must
      be prepared in hexane or iso-octane.  Each solution must be prepared
      every six months, or sooner if the solution has degraded or evaporated.

 4.    Gas Chromatograph Operating Conditions

      The following are the gas chromatographic analytical conditions.   The
      conditions are recommended unless otherwise noted.

          Carrier Gas:                     Helium
                                          (Hydrogen may be used, see 2.3)

          Column Flow:                     5 mL/min

          Make-up Gas:                     P-5/P-10 or N2 (required)
          Injector Temperature:            > 200 "C (see 4.2)

          Injection:                       On-column

          Injection Volume:                1 or 2 uL (see 4.1)

          Injector:                        Grob-type, splitless

          Initial Temperature:             150°C

          Initial Hold  Time:               1/2 min

          Temperature Ramp:                5°C to 6°C/min

          Final Temperature:               275°C

          Final Hold Time:                 Until after Decachlorobiphenyl has
                                          eluted (approximately 10 minutes)

      Optimize GC conditions for analyte separation and sensitivity.  Once
      optimized, the same GC conditions must be used for the analysis of all
      standards, samples, blanks, and MS/MSDs.

      The linearity of the ECD may be greatly dependent on the flow rate of
      the make-up gas.   Care must be taken to maintain stable and appropriate
      flow of make-up gas to the detector.

4.1   Manual injections must be 2.0 uL.  Auto injectors may use 1.0 uL
      volumes.   The same injection volume must be used for all standards,
      blanks,  and samples.

4.2   Cold (ambient temperature) on-column injectors that allow injection
      directly onto a 0.53 mm ID column may be used as long as the acceptance
      criteria for resolution,  calibration, and analyte breakdown are met.
                                 D-39/PEST                       OLM01.2  1/91

-------
5.

5.1
Analysis Sequence for Standards and Samples

All acceptable samples must be analyzed within a valid analysis
sequence as given below.
Time


0 hr.




Injection #
1 - 15

16

17

18
Material Injected
First 15 steps of the
Initial Calibration
Instrument Blank at end of
Initial Calibration
PEM at end of
Initial Calibration
First Sample
      12 hr.
Another 12 hr.
Another 12 hr.
                 o
                 o
           1st injection
             past 12:00  hr.
           2nd and 3rd
             injections
              past 12:00 hr.
           o
           o
           o
           o
           o
           o
           1st injection
             past 12:00  hr.
           2nd injection
           o
           o
           o
           o
           o
           o
           1st injection
             past 12:00  hr.
           2nd and 3rd
             injections
              past 12  hr.
           o
           o
           o
           o
           o
           etc.
Subsequent Samples

Last sample
Instrument Blank

Individual Standard Mixtures A and B
                                        Sample

                                        Subsequent Samples
Last Sample
Instrument Blank

Performance Evaluation Mixture
Sample

Subsequent Samples
Last Sample
Instrument Blank

Individual Standard Mixtures A and B
                                        Sample

                                        Subsequent Samples
      NOTE:   The first 12 hours are counted from the injection #16 (the
      Instrument Blank at the end of the initial calibration sequence), not
      from injection #1.   Samples may be injected until 12:00 hours have
      elapsed.   All subsequent 12-hour periods are timed from the injection
                                D-40/PEST
                                                               OLM01.0

-------
                                                             SECTION III
      of the instrument blank that bracketts the front end of the samples.
      Because the 12-hour time period is timed from injection of the
      instrument blank until the injection of the last sample, each 12-hour
      period may be separated by the length of one chromatographic run,  that
      of the analysis of the last sample.   While the 12-hour period may not
      be exceeded, the laboratory may run instrument blanks and standards
      more frequently, for instance to accomodate .staff working on 8-hour
      shifts.

5.2   Before any samples are analyzed, it is necessary for the Contractor to
      complete an acceptable initial calibration sequence (see paragraph 6).

5.3   After the initial calibration, the analysis sequence may continue as
      long as acceptable instrument blanks, Performance Evaluation Mixtures,
      and Individual Standard Mixtures A and B are analyzed at the required
      frequency (see paragraph 7).  This analysis sequence shows only the
      minimum required blanks and standards.  More blanks and standards may
      be run at the discretion of the Contractor; these must also satisfy the
      criteria presented in paragraph 7 in order to continue the run
      sequence.

5.4   An analysis sequence must also include all required matrix spike/matrix
      spike duplicate analyses and method blanks, but the Contractor may
      decide at what point in the sequence they are to be analyzed.

5.5   A standard of any identified Aroclor must be run within 72 hours of its
      detection in a sample chromatogram.

6..    Initial Calibration

6.1   Initial Calibration Sequence

      6.1.1   Before any samples  are  analyzed, it  is necessary for  the
              Contractor to complete  the  initial calibration  sequence given
              below.

              NOTE:  Steps 16 and 17  are  used as part of  the  calibration
              verification as well  (see paragraph  7).
                                D-41/PEST                             OLM01.0

-------
                         INITIAL CALIBRATION SEQUENCE

                              1.  Resolution Check
                              2.  Performance Evaluation Mixture
                              3.  Aroclor 1016/1260
                              4.  Aroclor 1221
                              5.  Aroclor 1232
                              6.  Aroclor 1242
                              7.  Aroclor 1248
                              6.  Aroclor 1254
                              9.  Toxaphene
                             10.  Low Point Standard A
                             11.  Low Point Standard B
                             12.  Midpoint Standard A
                             13.  Midpoint Standard B
                             14.  High Point Standard A
                             15.  High Point Standard B
                             16.  Instrument Blank
                             17.  Performance Evaluation Mixture

      6.1.2    Samples may be analyzed only after the initial calibration
               acceptance criteria (6.2) are met.  Otherwise, the analytical
               system is not functioning adequately for use with this
               protocol.

      6.1.3    The  initial calibration may continue to be used as long as the
               analytical system remains under control.  The proof that the
               analytical system is under control is provided by the analyses
               of the Performance Evaluation Mixtures.  If those analyses do
               not  meet the criteria described in paragraph 7, appropriate
               corrective action must be taken, and the initial calibration
               sequence must be  repeated.  The calibration sequence must also
               be repeated if any major change in instrument hardware or
               instrument parameters is made  (e.g., if a new column is
               installed or if the detector temperature is changed).

6.2   Initial Calibration Acceptance Criteria (apply to each GC column
      independently)

      6.2.1    The  initial calibration sequence must be analyzed in the order
               listed in paragraph 6.1 using  the optimized GC/EC operating
               conditions described  in paragraph 4.  The standards must be
               prepared according to paragraph 3.  Calculate the calibration
               factors and retention times according  to paragraphs 8-10.

      6.2.2    The  resolution criterion is that the depth of the valley
               between two adjacent peaks in  the Resolution Check Mixture must
               be greater than or equal to 60.0% of the height of the shorter
               peak.  The poorest resolution  on the DB-608 column probably
               will be between DDE and Dieldrin, between Methoxychlor and
               Endrin ketone and between Endosulfan I and gamma-Chlordane.  On
               the  DB-1701 column, resolution difficulties most  frequently
               occur between Endosulfan I and gamma-Chlordane, and between
               Methoxychlor and  Endosulfan sulfate.

                                D-42/PEST                       OLM01.3   2/91

-------
6.2.3    The breakdown of DDT and Endrin in both  of  the Performance
         Evaluation Mixtures  must be  less than 20.0  percent, and the
         combined breakdown of DDT and Endrin  must be  less  than 30.0
         percent where,

                                                                    EQ.2
% Breakdown DDT - Amount found in ng  (DDD+DDE) * 100
                     Amount  in ng of DDT injected

                                                                    EQ.3
% Breakdown Endrin'-
Amount found in ng  (Endrin aldehyde + Endrin ketone) * 100
              Amount of Endrin injected in ng

                                                                    EQ.4
Combined % Breakdown - %Breakdown DDT + %Breakdown Endrin

6.2.4    All  peaks  in  both  of the Performance  Evaluation Mixtures must
         be  100  percent resolved  on both columns.

6.2.5    The  absolute  retention times  of each  of  the single component
         pesticides  and surrogates  in  both  of  the PEMs must be within
         the  retention time windows determined from  the three-point
         initial  calibration,  in  paragraph  8.4.

6.2.6    The  relative  percent difference  of the calculated  amount and
         the  true  amount  for  each of the  single component pesticides and
         surrogates  in both of the  PEMs must be less than or equivalent
         to  25.0 percent, using equation  5.

6.2.7    At  least one  chromatogram  from  each of the  two Individual
         Standard Mixtures A  and  B, run  during the initial  calibration,
         must yield  peaks that give recorder deflections of 50 to 100
         percent of  full  scale.

6.2.8    The  resolution between any two  adjacent peaks in the midpoint
         concentrations of  Individual  Standard Mixtures A and B in the
         initial calibration  must be greater than or equal  to 90.0
         percent.

6.2.9    The  % RSD of  the calibration  factors  for each single component
         target compound must be  less  than  or  equal  to 20.0 percent,
         except as noted below.   The % RSD  of  the calibration factors
         for  the two surrogates must be  less than or equal  to 30.0
         percent.  Up  to  two  single component  target compounds (but not
         surrogates) per column may exceed  the 20.0 percent limit for
         %RSD, but  those compounds must have a % RSD of less than or
         equal to 30.0 percent.

         %RSD = Standard Deviation  x  100
                     Mean
                          D-43/PEST                      OLM01.5  4/91

-------
                                                             SECTION III
               Where,
               Standard Deviation
                                        n
                                            ci - x)'
                                             n-1
1/2
              Where,

              Xi - each individual value used to calculate the mean

              x  - the mean of n values

              n  = the total number of values

6.3   Corrective Action

      6.3.1   If the  technical acceptance criteria for the initial
              calibration are not met, inspect the system for problems.  It
              may be  necessary to change the column, bake out the detector,
              clean the injection port, or take other corrective actions to
              achieve the acceptance criteria.

      6.3.2   Contamination should be suspected as a cause if the detector
              cannot  achieve acceptable linearity using this method.  In the
              case of light contamination, baking out the detector at an
              elevated temperature (350°C) should be sufficient to achieve
              acceptable performance.  In the case of heavy contamination,
              passing hydrogen through the detector 1-2 hours at an elevated
              temperature may correct the problem.  In the case of severe
              contamination, the detector may require servicing by the ECD
              manufacturer.  DO NOT OPEN THE DETECTOR.  THE ECD CONTAINS
              RADIOCHEMICAL SOURCES.

      6.3.3   If a laboratory cleans out a detector using an elevated
              temperature, the ECD electronics must be turned off during the
              bake out procedure.

      6.3.4   After bake out or hydrogen reduction, the detector must be
              recalibrated using the initial calibration sequence.

      6.3.5   Initial calibration technical acceptance criteria MUST be met
              before  any samples or required blanks are analyzed.  Any
              samples or required blanks analyzed after the initial
              calibration criteria have not been met will require reanalysis
              at no additional cost to the Agency.

7.     Calibration Verification

7.1   Three types of analyses  are  used to  verify the calibration and evaluate
instrument performance.   The analyses of instrument blanks,  Performance
Evaluation Mixtures  (PEM),  and the mid point  concentration of Individual
Standard Mixtures A  and B constitute  the continuing
                                D-44/PEST                      OLM01.2   1/91

-------
      calibration.  Sample data are not acceptable unless bracketed by
      acceptable analyses of instrument blanks,  PEM,  and both Individual
      Standard Mixtures A and B.

7.2   An instrument blank and the Performance Evaluation Mixture must bracket
      one end of a 12-hour period during which sample data are collected,  and
      a second instrument blank and the the mid paint concentration of
      Individual Standard Mixtures A and B must bracket the other end of the
      12-hour period.

7.3   For the 12-hour period immediately following the initial calibration
      sequence, the instrument blank and the PEM that are the last two steps
      in the initial calibration sequence bracket the front end of that 12-
      hour period.  The injection of the instrument blank starts the
      beginning of that 12-hour period (see paragraph 5.1).  Samples may be
      injected for 12 hours from the injection of the instrument blank.   The
      three injections immediately after that 12-hour period must be an
      instrument blank, Individual Standard Mixture A, and Individual
      Standard Mixture B.  The instrument blank must be analyzed first,
      before either standard.  The Individual Standard Mixtures may be
      analyzed in either order (A,B or B,A).

7.4   The analyses of the instrument blank and Individual Standard Mixtures  A
      and B immediately following one 12-hour period may be used to begin the
      subsequent 12-hour period, provided that they meet the acceptance
      criteria in paragraphs 7.8-7.14.  In that instance, the subsequent 12-
      hour period must be bracketed by the acceptable analyses of an
      instrument blank and.a PEM, in that order.  Those two analyses may in
      turn be used to bracket the front end of yet another 12-hour period.
      This progression may continue every 12 hours until such time as any of
      the instrument blanks, PEMs, or Individual Standard Mixtures fails to
      meet the acceptance criteria in paragraphs 7.8-7.14.  The 12-hour time
      period begins with the injection of the instrument blank.  Standards
      (PEM or Individual Standard Mixtures), samples and required blanks may
      be injected for 12:00 hours from the time of injection of the
      instrument blank.

7.5   If more than 12 hours have elapsed since the injection of the
      instrument blank that bracketed a previous 12-hour period, an
      acceptable instrument blank and PEM must be analyzed in order to start
      a new sequence.  This requirement applies even if no analyses were
      performed since that standard(s) was injected.

7.6   After a break in sample analyses, the laboratory may only resume the
      analysis of samples using the current initial calibration for
      quantitation by analyzing an acceptable instrument blank and a PEM.

7.7   If the entire 12-hour period is not required for the analyses of all
      samples to be reported and all data collection is to be stopped, the
      incomplete sequence must be ended with either the instrument blank/PEM
      combination or the instrument blank/Individual Standard Mixtures A and
      B combination, whichever was due to be performed at the end of 12-hour
      period.
                                D-45/PEST                             OLM01.0

-------
7.8   Analysts are cautioned that running an instrument blank and a
      performance evaluation mixture once every 12 hours is the minimum
      contract requirement.   Late eluting peaks may carry over from one
      injection to the next if highly complex samples are analyzed or if the
      GC conditions are unstable.  Such carryover is unacceptable.
      Therefore, it may be necessary to run instrument blanks and performance
      evaluation mixtures more often to avoid discarding data.

7.9   The requirements for running the instrument blanks, Performance
      Evaluation Mixture, and Individual Standard Mixtures A and fi are waived
      when no samples , method blanks ,  or matrix spikes are run during that
      12-hour period.  After a break in sample analysis,  a laboratory may
      resume the analysis of samples,  method blanks, and matrix spikes and
      may use the current initial calibration for quantitation only after an
      acceptable PEM is run (paragraphs 7.2 - 7.6).   If a successful PEM
      cannot be run after an interruption,  an acceptable initial calibration
      must be run before sample data may be collected.  All acceptable sample
      analyses must be bracketed by acceptable performance evaluation
      mixtures and instrument blanks.

7.10  Technical Acceptance Criteria (apply to each GC column independently)

      7.10.1  All single component pesticides and surrogates  in the
              Performance Evaluation Mixtures used  to demonstrate continuing
              calibration must be 100 percent resolved.  The  resolution
              between any two adjacent peaks in the midpoint  concentrations
              of Individual Standard Mixtures A and B in the  initial
              calibration must be greater  than or equal  to 90.0 percent.

      7.10.2  The absolute retention time  for each of the single component
              pesticides and surrogates  in the PEMs and mid point
              concentration of the Individual Standard Mixtures used  to
              demonstrate continuing calibration must be within the retention
              time window determined from  the three-point  initial calibration
              in paragraph 8.4.

      7.10.3  The relative percent difference of the  calculated amount and
              the true amount for each of  the single  component pesticides and
              surrogates in the  PEM and  mid  point concentration of  the
              Individual Standard Mixtures used to  demonstrate continuing
              calibration must be less than  or equal  to  25.0  percent, using'-
              Equation 5.

                I cnom "  ccalc I
           RPD --    x 100                                  EQ.  5
              Cnom  =  true concentration of each analyte

              ^calc =  calculated concentration of each analyte from  the
              analyses of the standard
                                D-46/PEST                       OLM01.3   2/91

-------
                                                             SECTION III
              Note:   The vertical bars  in the equation indicate the absolute
              value,  hence RPD  is always a positive number.

      7.10.4  The percent breakdown of  DDT and Endrin in the PEM must be less
              than or equal  to  20.0 percent on both columns.  The combined
              breakdown of DDT  and Endrin must be less than or equal to 30.0
              percent on both columns.

      7.10.5  All instrument blanks must meet the acceptance criteria in
              paragraph 15.3.

7.11  Corrective Action

      7.11.1  If the  technical  acceptance criteria for the calibration
              verification are  not met, inspect the system for problems and
              take corrective action  to achieve the acceptance criteria.

      7.11.2  Major corrective  actions  such as replacing the GC column or
              baking  out the detector will require that a new initial
              calibration be performed  and meets the technical acceptance
              criteria in 6.2.

      7.11.3  Minor corrective  actions  may not require performing a new
              initial calibration, provided that a new analysis of the
              standard (PEM  or  Individual Mixture) that originally failed the
              criteria and an associated instrument blank immediately after
              the corrective action do  meet all the acceptance criteria.

      7.11.4  If the  analysis of the standard and instrument blank in 7.11.3
              fail any of the technical acceptance criteria, a new initial
              calibration must  be performed.

8.     Determination of Absolute Retention Times.

8.1   During the initial calibration sequence,  absolute retention times (RT)
      are determined for all single response pesticides,  the surrogates,  and
      at least three major peaks of each multicomponent analyte.

8.2   For single component pesticides, an RT is measured in each of three
      calibration standards and the mean RT is calculated as the average of
      the three values.  An RT  is measured for the surrogates in each of the
      three analyses of Individual Mixture A during the initial calibration
      and the mean RT is calculated as the average of the three values.

8.3   A retention time window is calculated for each single component analyte
      and surrogate by using the list in paragraph 8.4.   Windows are centered
      around the mean absolute retention time for the analyte established
      during the initial calibrations.
                                 D-47/PEST                       OLM01.2  1/91

-------
8.4   Retention time windows for single and multicomponent analytes and
      surrogates.
                                          Retention Time Window
                    Compound                    in Minutes

                  alpha-BHC                        +0.05
                  beta-BHC                         ±0.05
                  gamma-BHC                        ±0.05
                  delta-BHC                        ±0.05
                  Heptachlor                       ±0.05
                  Aldrin                           ±0.05
                  alpha-Chlorodane                 ±0.07
                  gamma-Chlorodane                 ±0.07
                  Heptachlor epoxide               ±0.07
                  Dieldrin                         ±0.07
                  Endrin                           ±0.07
                  Endrin aldehyde                  ±0.07
                  Endrin ketone                    ± 0.07
                  ODD                              ±0.07
                  DDE                              ±0.07
                  DDT                              ±0.07
                  Endosulfan I                     ±0.07
                  Endosulfan II                    ±0.07
                  Endosulfan sulfate               ±0.07
                  Methoxychlor                     ±0.07
                  Aroclors                         ±0.07
                  Toxaphene                        ±0.07
                  Tetrachloro-m-xylene             ±0.05
                  Decachlorobiphenyl               ±0.10

8.5   For each multicomponent analyte,  the RTs for three to five peaks are
      calculated from the initial calibration standard analysis.  An RT
      window of ±0.07 minutes is used for all multicomponent analyte peaks.

8.6   Analytes are identified when peaks are observed in the RT window for
      the compound on both GC columns.

9.     Calibration Factors for Single Component Pesticides.

9.1   During the initial calibration sequence, the Contractor must establish
      the magnitude of the linear BCD response range for each single
      component pesticide and surrogate on each column and for each GC
      system.  This is accomplished by analyzing the Individual Standard
      Mixtures A and B at three concentrations during the initial calibration
      sequence in paragraph 6.

9.2   The linearity of the instrument is determined by calculating a percent
      relative standard deviation (%RSD) of the calibration factors from a
      three-point calibration curve for each single component pesticide and
      surrogate.   Either peak area or peak height may be used to calculate
      calibration factors used in the %RSD equation.  For example, it is
      permitted to calculate linearity for Endrin based on peak area and to
      calculate linearity for Aldrin based on peak height.  It is not


                                D-48/PEST                      OLM01.1  12/90

-------
 permitted within a  %RSD calculation for an analyte to use calibration
 factors  calculated  from both peak area and peak height.  For example,
 it  is not permitted to calculate the calibration factor for the low
 point standard  for  endrin using peak height and calculate the midpoint
 and high point  standard calibration factors for endrin using peak area.

 9.2.1    Calculate  the calibration  factor  for  each  single component
         pesticide  and surrogate  over  the  initial calibration range
         using Equation 6.   The calibration factors for the  surrogates
         are calculated from the  three analyses  of  Individual Standard
         Mixture A  only.

 9.2.2    Calculate  the mean and the %RSD of the  calibration  factors for
         each single component  pesticide and surrogate  over  the  initial
         calibration range  using  Equations 7 and 8.

         CF =   Peak Area  (or Height) of  the Standard             EQ. 6
                         Mass  Injected (ng)

         	      n       CFi
         CF  -   I       	                                     EQ.  7
               i-1      n

         % RSD  - Standard  Deviation
                        7Z            x  10°
                        CF                                        EQ.  8


         CF  = Calibration factor


         Where,
        Standard  Deviation  -
                                  n
                                        -x)2
                                       n-1
1/2
        Where,

        xi — each individual value used to calculate the mean

        x  =  the mean of n values

        n  =  the total number of values

9.2.3   The linearity of the calibration  is considered acceptable when
        the % RSD of the three point calibration is less than  20.0
        percent except as noted in the following.

        The % RSD of the two surrogates must be less than or equal  to
        30.0 percent.  Up to two single component target compounds  (but
        not surrogates) per column may exceed  the 20.0 percent limit
        for % RSD., but those compounds must have a % RSD of less than
        or equal to 30.0 percent.


                          D-49/PEST                      OLM01.5  4/91

-------
      9.2.4    If the  linearity  requirements listed above are met, the
               calibration  factor  from  the mid point concentration standard is
               used  for  quantitation of each single component pesticide.

9.3   Sample analysis may not proceed until a satisfactory calibration has
      been demonstrated.

10.   Calibration Factors for Toxaphene and Aroclors

10.1  Toxaphene and Aroclors require only a single-point calibration and they
      present special analytical difficulties.   Because of the alteration of
      these materials in the environment, it is probable that samples which
      contain multicomponent analytes will give patterns similar to, but not
      identical with, those of the standards.

10.2  A set of three to five major peaks  is selected for each multicomponent
      analyte.  Retention times (see 8.4) and calibration factors are
      determined from the initial calibration analysis for each peak.
      Guidance for the choice of which peaks to use is given in paragraph
      13.9

11.   Acceptance Criteria for Chromatograms of Calibration Standards

      The identification of single component pesticides by gas
      chromatographic methods is based primarily on rentention time data.
      The retention time of the apex of a peak can be verified only from an
      on-scale chromatogram.  The identification of multicomponent analytes
      is based primarily on recognition of patterns of retention times
      displayed on a chromatogram.  Therefore,  the following requirements
      apply to all data presented for single component and multicomponent
      analytes.

11.1  The chromatograms that result from the analyses of the Resolution Check
      Mixture, the Performance Evaluation Mixture, and Individual Standard
      Mixtures A and B during the initial calibration sequence must display
      the single component analytes present in each standard at greater than
      10 percent of full scale but less than 100 percent of full scale.

11.2  The chromatograms, for at least one of the three analyses each of
      Individual Standard Mixtures A and B from the initial calibration
      sequence, must display the single component analytes at greater than 50
      percent and less than 100 percent of full scale.

11.3  The chromatograms of the standards  for the multicomponent analytes
      analyzed during the initial calibration sequence must display the peaks
      chosen for identification of each analyte at greater than 25 percent
      and less than 100 percent of full scale.

11.4  For any standard containing alpha-BHC,  the baseline of the chromatogram
      must return to below 50 percent of full scale before the elution time
      of alpha-BHC,  and return to below 25 percent of full scale after the
      elution time of alpha-BHC and before the elution time of
      decachlorobiphenyl.
                                D-50/PEST                             OLM01.0

-------
 11.5  If a chromatogram is replotted electronically to meet requirements, the
      scaling factor used must be displayed on the chromatogram.

 11.6  If the chromatogram of any standard needs to be replotted
      electronically to meet these requirements,  both the initial
      chromatogram and the replotted chromatogram must to submitted in the
      data package.

 12.   Sample Analysis

 12.1  Unless ambient temperature on-column injection is used (see paragraph
      4.2), the injector must be heated to at least 200°C.  The optimized gas
      chromatographic conditions from paragraph 4 must be used.

 12.2  The injection must be made on-column by using either automatic or
      manual injection.  If autoinjectors are used, 1.0 uL injection volumes
      may be used.  Manual injections shall use at least 2.0 uL injection
      volumes.  The same injection volume must be used for all standards,
      samples, and blanks associated with the same initial calibration.  If a
      single injection is used for two GC columns attached to a single
      injection port, it may be necessary to use an injection volume greater
      than 2 uL.  However, the same injection volume must be used for all
      analyses.

 12.3  Analysis of a sample on both GC columns is required for all samples,
      blanks,  matrix spikes, and matrix spike duplicates.

 12.4  The requirements for the analysis sequence apply to both GC columns and
      for all instruments used for these analyses.

 12.5  The laboratory will identify and quantitate analyte peaks based on RT
      and calibration factor established during the initial calibration
      sequence,  as long as an acceptable calibration verification (see
      paragraph 7) is performed every 12 hours.

 12.6  The protocol is intended to achieve the quantitation limits shown in
      Exhibit C whenever possible.  If sample chromatograms have interfering
      peaks,  a high baseline, or off-scale peaks, then those samples must be
      reanalyzed following dilution, further cleanup, or reextraction.
      Samples which cannot be made to meet the given specifications after one
      reextraction and three-step cleanup (GPC, Florisil, and sulfur removal)
      are reported in the SDG Narrative and do not require further analysis.
      No limit is placed on the number of reextractions of samples that may
      be required because of contaminated method blanks.

12.7  The sample must be analyzed at the most concentrated level that is
      consistent with achieving satisfactory chromatography (defined below).
      If dilution is employed solely to bring a peak within the calibration
      range or to get a multicomponent pattern on scale,  the results for both
      the more and the less concentrated extract must be reported.  The
      resulting changes in quantitation limits and surrogate recovery must be
      reported also for the diluted samples.
                                D-51/PEST                       OLM01.5  4/91

-------
12.8  If the Contractor has reason to believe  that  diluting the  final
      extracts will be necessary,  an undiluted run  may not  be  required.   If
      an acceptable chromatogram (as defined below)  is achieved  with the
      diluted extract, an additional extract 10 times  the concentration of
      the dilute sample must be injected and reported  with  the sample  data.

12.9  No target analyte concentrations may exceed .the  upper limit of the
      initial calibration.

12.10 A standard for any identified multicomponent  analyte  must  be analyzed
      during a valid analytical sequence on the same instrument, within 72
      hours of its detection in a sample.

12.11 The identification of single component pesticides by  gas
      chromatographic methods is based primarily on retention  time data.   The
      retention time of the apex of a peak can be verified  only  from an on-
      scale chromatogram.  The identification  of multicomponent  analytes  is
      based primarily on recognition of patterns of retention  times displayed
      on a chromatogram.  Therefore, the following  requirements  apply  to  all
      data presented for single component and  multicomponent analytes.

      12.11.1 When no analytes are  identified in a sample,  the chromatograms
              from the  analyses of  the sample extract must use the same
              scaling factor as was used for the low point standard of the
              initial calibration associated with those analyses.

      12.11.2 Chromatograms must display single component pesticides detected
              in  the sample at less than full scale.

      12.11.3 Chromatograms must display the largest peak of any
              multicomponent analyte detected in the sample at  less than  full
              scale.

      12.11.4 If  an extract must be diluted, chromatograms must display
              single component pesticides between 10 and 100 percent of full
              scale.

      12.11.5 If  an extract must be diluted, chromatograms must display the
              peaks chosen for quantitation of multicomponent analytes
              between 25 and 100 percent of full scale.

      12.11.6 For any sample, the baseline of the chromatogram must return  to
              below 50  percent of full scale before the elution time of
              alpha-BHC, and return to below 25 percent of full scale after
              the elution time of alpha-BHC and before the elution time of
              decachlorobiphenyl.

      12.11.7 If  a chromatogram is  replotted electronically to  meet these
              requirements, the scaling factor used must be displayed on  the
              chromatogram.
                                D-52/PEST                      OLM01.2   1/91

-------
                                                             SECTION III
      12.11.8 If the chromatogram of any sample needs to be replotted
              electronically to meet these requirements, both the initial
              chromatogram and the replotted chromatogram must be submitted
              in the data package.

13.   Ouantitation of Analvtes

13.1  Quantitation of target analytes and surrogates  must be performed and
      reported on both columns.

13.2  Analytes must be quantitated with an electronic integrator or  with a
      laboratory data system.   The analyst can use either peak height or peak
      area as the basis for quantitation.   The use of an electronic
      integrator or a laboratory data system is required.

13.3  The chromatograms of all  samples must be reviewed by a qualified
      pesticide analyst before  they are reported.

13.4  In order to be quantitated,  the detector response (peak area or peak
      height) of all of the single component analytes must lie between the
      response of the low and high concentrations  in  the initial calibration.
      If the analytes are detected below the CRQL,  they are reported as
      present below the CRQL, and flagged according to the instructions in
      Exhibit B.  If they are detected at a level  greater than the high
      calibration point, the sample must be diluted either to a  maximum of
      1:100,000 or until the response is within the linear range established
      during calibration.  Guidance in performing  dilutions and  exceptions to
      this requirement are given below.

      13.4.1  If the response is still above the high calibration point after
              the dilution of 1:100,000, the Contractor shall contact the SMO
              immediately.

      13.4.2  Use the results of the original analysis to determine the
              approximate dilution factor required to get the largest analyte
              peak within the initial calibration range.

      13.4.3  The dilution factor chosen should keep the response of the
              largest peak for a target compound in the upper half of the
              initial calibration range of the instrument.

      13.4.4  Do not submit data for more than two analyses,  i.e., the
              original  sample extract and one dilution, or, if a screening
              procedure was employed, from the most concentrated dilution
              analyzed  and one further dilution.

      13.4.5  Do not dilute MS/MSD samples to get either spiked or non-spiked
              analytes within the calibration range.   If the  sample from
              which the MS/MSD aliquots were taken contains high levels of
              the spiked analytes, calculate the concentration and recovery
              of the analytes from the undiluted analysis and note the
              problem in the SDG Narrative.
                                D-53/PEST                      OLM01.2   1/91

-------
13.5  The concentrations of the single  component pesticides  are  calculated
      separately for both GC columns  by using  the  following  equations:

      13.5.1  Water
                                        (Ax)(Vt)(Df)                   EQ. 9
              Concentration    ug/L  -
                                        (CF)(V0)(Vi)

              Where
              Ax  =•  Area of the peak for the compound to be measured

              CF  -  Calibration factor for the mid point concentration
                     external standard (area per ng)

              Vo  -  Volume of water extracted in milliliters (mL)

              V^  -  Volume of extract injected in microliters (uL)
                     (If a single injection is made onto two columns, use one
                     half the volume in the syringe as the volume injected
                     onto to each column.)
              Vt  -  Volume of the concentrated extract in raicroliters (uL)
                     (this volume must be 10000 uL, see Section II, 7.2.3)

              Df  =  Dilution Factor.  The dilution factor for analysis of
                     water samples by this method is defined as follows:

              uL most cone,  extract used to make dilution + uL clean solvent
                        uL most cone, extract used to make dilution

              If no dilution is performed, Df - 1.0.

              If GPC is performed on a water sample extract,  Vt becomes 5000
              uL, and a factor of 2 must be added to the numerator,  as
              described below for soil/sediment samples.

      13.5.2   Soil/Sediment

                                           (Ax)(Vt)(Df)(2.0)           EQ.  10
              Concentration     ug/Kg  =  	
              (Dry weight basis)          (CF) (Vj.) (Ws) (D)

              Where
              Ax and CF are as given for water,  above.

              Vt  =  Volume of the concentrated extract  in microliters  (uL)
                     (this volume must be 5000 uL, see Section II, 7.2.3)

              V^  =  Volume of extract injected in microliters (uL)
                     (If a single injection is made onto two columns, use one
                     half the volume in the syringe as the volume injected
                     onto to each column.)

              D   =  100 - % moisture
                         100

              Ws  =  Weight of sample extracted in grams (g)
                                D-54/PEST                      OLM01.3  2/91

-------
                                                       SECTION III

         Df  -  Dilution Factor.   The dilution factor for analysis of
                soil samples by this method is defined as follows:
         uL most cone,  extract used to make  dilution  + uL clean solvent
                  uL most cone, extract used to make  dilution

         If no dilution is performed, Df -  1.0.

         The factor of  2.0 in the  numerator  is used to account for the
         amount of extract that is not recovered  from the mandatory use
         of GPC cleanup.   Concentrating the  extract collected after GPC
         to 5.0 mL rather than 10.0 mL for water  samples  not subjected
        •to GPC (see Section II, 7.2.3), maintains the sensitivity of
         the soil method comparable to that  of the water  method,  but
         correction of  the numerical result  is still  required.

13.5.3   Note that the  calibration factors used for the quantitation of
         the single component pesticides are the  calibration factors
         from the mid point concentration standard for each analyte.

13.5.4   Because of the likelihood that compounds co-eluting with the
         target compounds will cause positive interferences and increase
         the concentration determined by the method,  the  lower of the
         two concentrations calculated for  each single component
         pesticide is reported on  Form I.   In addition, the
         concentrations calculated for both  the GC columns are reported
         on Form X,  along with a percent difference comparing the two
         concentrations.   The percent difference  is calculated according
         to Equation 11.

                                                                  EQ. 11

         %D = ConcH  - ConcL   x 100
Where,

Concn = The higher of the two concentrations for the target compound in
         question

ConcL = The lower of the two concentrations for the target compound in
         question

Note that using  this equation will result in percent difference values
that are always  positive.  The value will also be greater than a value
calculated using the higher concentration in the denominator, however,
given the likelihood of a positive interference raising the
concentration determined on one GC column, this is a conservative
approach to comparing the two concentrations.
                           D-55/PEST                       OLM01.3  2/91

-------
                                                             SECTION III


3.6   The concentrations of the surrogates are calculated separately for both
      GC columns in a similar manner as the other analytes,  using Equations 9
      and 10.  Use the calibration factors from the midpoint concentration of
      Individual Standard Mixture A.  The recoveries of the  surrogates are
      calculated for both GC columns acccording to Equation  12.

                                          Qd                      EQ.  12
            Surrogate Percent Recovery -  ^—  x  100
                                          Va

      Where,

      Qd = Quantity determined by analysis

      Qa = Quantity added to sample/blank

      The limits for the recovery of the surrogates are 60-150 percent for
      both surrogate compounds.  As these limits are only advisory,  no
      further action is required by the laboratory is required,  however,
      frequent failures to meet the limits for surrogate recovery warrant
      investigation by the laboratory, and may result in questions from the
      Agency.  Surrogate recovery data from both GC columns  are reported (see
      Exhibit B).

13.7  The quantitative determination of Toxaphene or Aroclors is somewhat
      different from that of single component pesticides. Quantitation of
      peaks within the detector linear range CRQL to > 16 times CRQL is based
      on a single calibration point assuming linear detector response.
      Alternatively, a linear calibration range may be established during a
      run sequence by a three-point calibration curve for any multicomponent
      analyte.  If the concentration is calculated to be 10   times the CRQL,
      the Contractor shall contact the SMO immediately.

13.8  The quantitation of Toxaphene or Aroclors must be accomplished by
      comparing the heights or the areas of each of the three to five major
      peaks of the multicomponent analyte in the sample with the calibration
      factor for the same peaks established during the initial calibration
      sequence.  The concentration of multicomponent analytes is calculated
      by using Equations 9 and 10, where Ax is the area for  each of the major
      peaks of the multicomponent analyte.  The concentration of each peak is
      determined and then a mean concentration for three to  five major peaks
      is determined on both columns.  The following table lists the number of
      potential quantitation peaks for each Aroclor and Toxaphene.

                                         No.  of  Potential
                Analyte	        Quantitation  Peaks

               Aroclor  1016/1260                 5/5
               Aroclor  1221                      3
               Aroclor  1232                      4
               Aroclor  1242                      5
               Aroclor  1248                      5
               Aroclor  1254                      5
               Toxaphene                         4
                                 D-56/PEST                      OLM01.3  2/91

-------
                                                             SECTION  III


13.9  The reporting requirements/for Toxaphene  and the  Aroclors are similar
      to those for the single component  analytes,  except that the lower mean
      concentration (from three to five  peaks)  is  reported on Form X,  and the
      two mean concentrations'are  compared using Equation 11.

13.10 The choice of the peaks used for multicomponent quantitation and the
      recognition of those peaks may be  complicated by  the environmental
      alteration of the Toxaphene  or Aroclors,  and by the presence of
      coeluting analytes or matrix interferences,  or both.

13.11 If more than one multicomponent analyte  is observed in a sample,  the
      Contractor must choose separate peaks to  quantitate the different
      multicomponent analytes.  A  peak common  to both analytes present in the
      sample must not be used to quantitate either compound.

14.   Sample Data Acceptance Criteria

14.1  The requirements below apply to both columns,  and quantitation must be
      performed on both GC columns and reported.

14.2  All samples must be be analyzed as part  of a valid analysis sequence
      (paragraph 5).  They must be bracketed by acceptable instrument blanks
      (paragraph 15.3), acceptable Performance  Evaluation Mixtures, and
      acceptable Individual Standard Mixtures  A and B (paragraph 7) that were
      analyzed at the required frequency.

14.3  The retention times for both of the surrogates must be within the
      retention time windows as calculated in  paragraph 8 for both GC
      columns.

14.4  Reportable data for a sample must  include a  chromatogram in which a
      baseline returns to below 50 percent of  full scale before the elution
      time of alpha-BHC, and to below 25 percent of full scale after alpha-
      BHC and before decachlorobiphenyl.

14.5  If dilution has been applied and if no peaks are  detected above 25
      percent of full scale, analysis of a more concentrated sample is
      required.

14.6  Reportable sample data must  include chromatogram(s) which meet the
      criteria in paragraph 12.11.

15.   Blanks

      There are two types of blanks always required by this method:  the
      method blank and the instrument blank.  A separate sulfur cleanup blank
      may be required if all samples associated with a given method blank are
      not subjected to sulfur cleanup.  Samples that are associated with a
      sulfur cleanup blank are also associated with the method blank with
      which they were extracted.  Both the method and sulfur cleanup blanks
      must meet the respective acceptance criteria for the sample analysis
      acceptance criteria to be met.
                                 D-57/PEST                       OLM01.3  2/91

-------
                                                             SECTION  III

15.1  Method blank

      15.1.1  Method blanks are spiked with the surrogate solution,
              extracted, cleaned up, and analyzed by following the same
              procedure that is used with the samples.   A water method blank
              is one liter of reagent water treated as the water sample
              aliquot.  A soil method blank is 30 g of sodium sulfate treated
              as the soil sample aliquot.

              Method blank analysis must be performed once for the following,
              whichever is most frequent, and analyzed on each GC/EC system
              used  to analyze samples:

              o   Each Case,  OR
              o   Each 14 calendar day period (7  calendar day period  for 14-
                  day data turnaround contracts)  during which samples in a
                  Case are received (said period  beginning with the receipt
                  of the first sample in that Sample Delivery Group), OR

              o   Each 20 samples  in a Case,  including  matrix spikes  and
                  reanalyses,  that are of similar matrix (water or soil),  OR

              o   Whenever samples are extracted  by the same procedure
                  (separatory funnel, continuous  liquid-liquid extraction,  or
                  sonication).
      15.1.2  In order to be acceptable, a method blank analysis cannot
              contain any of the analytes listed in Exhibit C at greater than
              the CRQL.  The surrogate retention times must be within the
              retention time windows calculated from the initial calibration
              sequence mean retention time for both tetrachloro-m-xylene and
              decachlorobiphenyl.

      15.1.3  All samples associated with an unacceptable method blank (see
              Form  IV) must be reextracted and reanalyzed at no additional
              cost  to the Agency.

15.2  Sulfur Cleanup Blank.

      15.2.1  The sulfur cleanup blank is a modified form of the method
              blank.  The sulfur cleanup blank is hexane spiked with  the
              surrogates and passed  through the sulfur cleanup procedure  (see
              Section II, paragraph  7.4).

      15.2.2  The sulfur cleanup blank is prepared when only part  of  a set of
              samples extracted together requires sulfur removal.  A  method
              blank is associated with the entire set of samples.  The sulfur
              cleanup blank is associated with the part of  the set which
              required sulfur cleanup.  If all the samples  associated with a
              given method blank are subjected to sulfur cleanup,  then the
              method blank must be subjected to sulfur cleanup, and  no
              separate sulfur cleanup blank is required.
                                 D-58/PEST                       OLM01.3  2/91

-------
                                                             SECTION  III

      15.2.3   In  order to be acceptable, a sulfur blank analysis cannot
               contain any of the analytes listed in Exhibit C at greater than
               the CRQL, assuming that the material in the sulfur blank
               resulted from the extraction of a 1 L water sample.  Calculate
               the concentration of each analyte using the equation in
               paragraph 13.5.1.  Compare the results to the CRQL values for
               water samples in Exhibit C.  The surrogate retention times must
               be  within the retention time windows calculated from the
               initial calibration sequence mean retention time for both
               tetrachloro-m-xylene and decachlorobiphenyl.

      15.2.4   All samples associated with an unacceptable sulfur blank (see
               Form IV) must be reextracted and reanalyzed at no additional
               cost to the Agency.

15.3  Instrument blank

      15.3.1   An  instrument blank is a hexane or iso-octane solution
               containing 20.0 ng/mL of tetrachloro-m-xylene and
               decachlorobiphenyl.

      15.3.2.  The first analysis in a 12-hour analysis sequence must be an
               instrument blank.  All acceptable samples analyses are to be
               bracketed by acceptable instrument blanks, as described in
               paragraph 5.1.

      15.3.3   An  acceptable instrument blank must be analyzed within a 12-
               hour analysis sequence and must demonstrate that no analyte in
               Exhibit C is detected at greater than 0.5 times the CRQL and
               that the surrogate retention times are within the retention
               time windows.  For comparing the results of the instrument
               blank analysis to the CRQLs, assume that the material in the
               instrument resulted from the extraction of a 1 L water sample
               and calculate the concentration of each analyte using the
               equation in paragraph 13.5.1.  Compare the results to one-half
               the CRQL values for water samples in Exhibit C.

      15.3.4 .If  analytes are detected at greater than half the CRQL, or the
               surrogate RTs are outside the RT windows, all data collection
               must be stopped, and corrective action must be taken.  Data for
               samples which were run between the last acceptable instrument
               blank and the unacceptable blank are considered suspect.  An
               acceptable instrument blank must be run before additional data
               are collected.  After an acceptable instrument blank is run,
               all samples which were run after the last acceptable instrument
               blank must be reinjected during a valid run sequence at no
               additional cost to the Agency and must be reported.

      15.3.5   Analysts are cautioned that running an instrument blank once
               every 12 hours is the minimum contract requirement.  Late
               eluting peaks may carry over from one injection to the next if
               highly complex samples are analyzed or if the GC conditions are
               unstable.  Such carryover is unacceptable.  Therefore,  it may
                                D-59/PEST                       OLM01.3  2/91

-------
                                                             SECTION III

              be  necessary  to  run  instrument blanks more often to avoid
              discarding  data.

16.   Matrix Spike/Matrix Spike Duplicate

16.1  A matrix spike and matrix spike duplicate must be extracted and
      analyzed at least once with every 20 samples of each matrix.   NOTE:
      There is no differentiation between "low" and "medium" soil samples in
      this method.  Therefore only one soil MS/USD is to be submitted per
      SDG.

16.2  The surrogate retention times must be within the retention time windows
      specified.

16.3  The percent recoveries and the relative percent difference between the
      recoveries of each of the 6 compounds in the matrix spike samples will
      be calculated and reported by using the following equations:

                                  SSR - SR
         Matrix Spike Recovery - 	  x 100
                                     SA                           Eq. 12

      Where,

          SSR  = Spike  sample  result
          SR  = Sample result
          SA  = Spike  added


              I MSR - MSDR I
       RPD	-	— x 100                                      Eq.  13
             1/2(MSR + MSDR)

      Where,
          RPD   =  Relative  percent difference
          MSR   =  Matrix spike  recovery
          MSDR =  Matrix spike  duplicate  recovery
      The vertical bars in the formula above indicate the absolute value of
      the difference, hence RPD is always expressed as a positive value.

16.4  The Contractor shall report matrix spike and matrix spike duplicate
      recoveries and percent difference values with the analytical results
      (see Exhibit B).   The limits for matrix spike compound recovery and RPD
      are given below.   As these limits are only advisory, no further action
      by the laboratory is required, however, frequent failures to meet the
      limits for recovery or RPD warrant investigation by the laboratory, and
      may result in questions from the Agency.
                                 D-60/PEST                       OLM01.3   2/91

-------
                                                             SECTION  III
                          MATRIX SPIKE RECOVERY AND
                      RELATIVE PERCENT DIFFERENCE LIMITS
Compound
gamma -BHC (Lindane)
Heptachlor
Aldrin
Dieldrin
Endrin
4, 4' -DDT
%Recovery
Water
56-123
40-131
40-120
52-126
56-121
38-127
RPD
Water
15
20
22
18
21
27
%Recovery
Soil
46-127
35-130
34-132
31-134
42-139
23-134
RPD
Soil
50
31
43
38
45
50
17.    GC/MS Confirmation of Pesticides and Aroclors

17.1  Any pesticide or Aroclor analyte listed in Ex.  C for which a
      concentration is reported from a GC/EC analysis must have the
      identification confirmed by GC/MS if the concentration is sufficient
      for that purpose.  The following paragraphs are to be used as  guidance
      in performing GC/MS confirmation.  If the the  Contractor fails to
      perform GC/MS confirmation as appropriate,  the Agency may require
      reanalysis of any affected samples at no additional cost to the Agency.

      17.1.1  The GC/MS confirmation may be accomplished by one of  three
              general means.

                         o   Examination of the  semivolatile GC/MS library
                             search results (i.e.  TIC data)
                         o   A second analysis  of the the  semivolatile
                             extract

                         o   Analysis of the pesticide/Aroclor extract,
                             following any solvent exchange  and concentration
                             steps  that may be  necessary.
      17.1.2  The semivolatile GC/MS analysis procedures outlined in  Ex.  D SV
              are based on  the injection  into the instrument of approximately
              20 ng of a target compound  in a 2 uL volume.  The semivolatile
              CRQL values in Ex. C are based on the  sample  concentration  that
              corresponds to an extract concentration of 10 ng/uL of  target
              analyte.  However, these are quantitation limits, and the
              detection of  analytes and generation of reproducible  mass
              spectra will  routinely be possible at  levels  3-10 times  lower.
              The sample concentration corresponding  to 10  ng/uL in extract
              will depend on the sample matrix.

                         o   For  water samples,  20 ng/ 2 uL corresponds to a
                             sample concentration of 10  ug/L.

                         o   For  soil samples prepared according to  the
                             semivolatile low level  soil method (i.  e. 30  g
                             of soil),  the corresponding sample concentration
                             is 330 ug/Kg.
                                D-61/PEST                      OLM01.3   2/91

-------
                                                       SECTION III

                   o   For soil samples  prepared according to  the
                       semivolatile  medium  level soil method (i. e.  1  g
                       of soil),  the corresponding  sample  concentration
                       is 10,000 ug/Kg.
        Therefore, based on the values  given above, any pesticide
        sample  in which  the sample  concentration is greater than or
        equal to an extract concentration of 10 ng/uL should enable the
        laboratory to confirm  the pesticide/Aroclor by GC/MS analysis
        of  the  semivolatile extract.

7.1.3   In  order to confirm the identification  of  the target
        pesticide/Aroclor, the laboratory must  also analyze a reference
        standard for the analyte.   In order to  demonstrate  the ability
        of  the  GC/MS system to identify the analyte in question, the
        concentration of the standard should be no greater  than 10
        ng/uL.

        17.1.3.1   To facilitate the confirmation of the
                   pesticide/Aroclor analytes from the semivolatile
                   library search data,  the laboratory may wish to
                   include these analytes in the semivolatile
                   continuing calibration standard at a concentration
                   of 10 ng/uL or less.   If added to this GC/MS
                   standard, the response factors,  retention times,
                   etc.  for these analytes would be reported on the
                   GC/MS quantitation report, but not on the GC/MS
                   calibration data reporting forms.  As only a single
                   concentration of each analyte would be analyzed, no
                   linearity (%RSD)  or percent difference criteria
                   would be applied to the response factors for these
                   additional analytes.

        17.1.3.2   The laboratory is advised that library search
                   results from the NIST/EPA/MSDC mass spectral library
                   will not likely list the name of the
                   pesticide/Aroclor analyte as it appears  in this SOW,
                   hence, the mass spectral interpretation  specialist
                   is advised to compare the CAS Registry numbers for
                   the pesticides/Aroclors to those from the library
                   search routine.

17.1.4  If  the  analyte cannot be confirmed from the semivolatile
        library search data for the original semivolatile GC/MS
        analysis, the laboratory may analyze another aliquot of the
        semivolatile sample extract after  further  concentration of  the
        aliquot.  This second  aliquot must either  be analyzed as  part
        of  a routine semivolatile GC/MS analysis,  including instrument
        performance checks (DFTPP), calibration standards containing
        the pesticides/Aroclors as  described in paragraph  17.1.3,  etc.,
        or  it must be analyzed along with  separate  reference  standards
        for the analytes  to be confirmed.
                          D-62/PEST                       OLM01.3  2/91

-------
                                                             SECTION III


      17.1.5   If the  analyte  cannot be confirmed by either  the procedures  in
               paragraphs  17.1.3  or 17.1.4,  then an aliquot  of the extract
               prepared for  the GC/EC  analysis must be analyzed by GC/MS,
               following any necessary solvent exchange and  concentration
               steps.   As  in paragraph 17.1.4, analysis of a reference
               standard is required if the GC/MS continuing  calibration
               standard does not  contain  the analyte to be confirmed.

      17.-1.6   Regardless  of which of  the three approaches above  is used for
               GC/MS confirmation, the appropriate blank must also be analyzed
               by GC/MS to demonstrate that  the presence of  the analyte was
               not the result  of  laboratory  contamination.   If the
               confirmation  is based "on the  analysis of the  semivolatile
               extract,  then the  semivolatile method blank extracted with the
               sample  must also be analyzed.  If the confirmation is based  on
               the analysis  of the extract prepared for the  GC/EC analysis,
               then the pesticide method  blank extracted with the sample must
               be analyzed.

17.2  If the identification of the analyte can not be confirmed by any of the
      GC/MS procedures above and the concentration calculated from the GC/EC
      analysis is greater than or equal to the concentration of the reference
      standard analyzed by GC/MS, then report the analyte as undetected,
      adjust the  sample quantitation limit  (the value associated with the "U"
      qualifier)  to a sample concentration equivalent to the concentration of
      the GC/MS  reference standard,  and qualify the result on Form I with one
      of the laboratory-defined  qualifiers  ("X",  "Y",  or "Z").   In this
      instance,  define the qualifier explicitly in the SDG Narrative,  and
      describe the steps  taken to confirm the analyte in the Narrative.

17.3  For GC/MS  confirmation of  single component analytes,  the required
      deliverables are c'opies of the library search results (three best TIC
      matches) or analyte spectrum and the spectrum of the  reference
      standard.   For multicomponent analytes,  spectra of three characteristic
      peaks are  required  for both the sample component and  the reference
      standard.

17.4  The purpose of GC/MS analysis is for confirmation of  identification,
not quantitation.  Therefore,  the concentrations of all pesticides/Aroclors
shall be based on the GC/EC results.   The exception noted in paragraph 17.2
applies only to  analytes  that cannot be confirmed at a concentration above
that of the reference standard.
                                 D-63/PEST                      OLM01.3  2/91

-------
                  EXHIBIT E
QUALITY ASSURANCE/QUALITY CONTROL REQUIREMENTS
                     E-l                               OLM01.0

-------
                               TABLE OF CONTENTS
SECTION
  PAGE
OVERVIEW  	   E-3




I     INTRODUCTION	   E-4




II    QUALITY ASSURANCE PLAN    	.'	   E-6




III   STANDARD OPERATING PROCEDURES    	   E-10




IV    QA/QC REQUIREMENTS




      Volatiles  (VGA) QA/QC Requirements    	   E-16/VOA




      Semivolatiles  (SV) QA/QC  Requirements    	   E-22/SV




      Pesticides/Aroclors  (PEST) QA/QC Requirements    	   E-28/PEST




V     ANALYTICAL STANDARDS REQUIREMENTS    	   E-36




VI    CONTRACT COMPLIANCE SCREENING    	   E-41




VII   REGIONAL DATA REVIEW    	   E-42




VIII  LABORATORY EVALUATION SAMPLES    	   E-43




IX    GC/MS TAPE AUDITS    	   E-46




X     DATA PACKAGE AUDITS   	   E-49




XI    ON-SITE LABORATORY EVALUATIONS   	   E-51




XII   QUALITY ASSURANCE AND DATA TREND ANALYSIS    	   E-54




XIII  DATA MANAGEMENT   	   E-55
                                     E-2
OLM01.0

-------
                                   OVERVIEW
Quality assurance and quality control are integral parts of the Environmental
Protection Agency's  (EPA) Contract Laboratory Program (CLP).  The quality
assurance (QA) process consists of management review and oversight at the
planning, implementation, and completion stages of the environmental data
collection activity, to ensure that data provided are of the quality
required. The quality control (QC) process includes those activities required
during data collection to produce the data quality desired and to document
the quality of the collected data.

During the planning  of an environmental data collection program, QA
activities focus on  defining data quality criteria and designing a QC system
to measure the quality of data being generated.  During the implementation of
the data collection  effort, QA activities ensure that the QC system is
functioning effectively, and that the deficiencies uncovered by the QC system
are corrected.  After environmental data are collected, QA activities focus
on assessing the quality of data obtained to determine its suitability to
support enforcement  or remedial decisions.

The purpose of this  Exhibit is to describe the overall quality
assurance/quality control operations and the processes by which the Program
meets the QA/QC objective defined above.  This contract requires a variety of
QA/QC activities. These contract requirements are the minimum QC operations
necessary to satisfy the analytical requirements associated with the
determination of the different method analytes.  These QC operations are
designed to facilitate laboratory comparison by providing the EPA with
comparable data from all Contractors.  These requirements do not release the
analytical Contractor from maintaining their own QC checks on method and
instrument performance.
                                     E-3                               OLM01.0

-------
                                   SECTION  I
                                 INTRODUCTION
Appropriate use of data generated under the large range of analytical
conditions encountered in environmental analyses requires reliance on the
quality control procedures and criteria incorporated into the methods.  The
methods in this contract have been validated on samples typical of those
received by the laboratories in the Contract Laboratory Program (CLP).
However the validation of these methods does not guarantee that they perform
equally well for all sample matrices encountered.  Inaccuracies can also
result from causes other than unanticipated matrix effects, such as sampling
artifacts, equipment malfunctions, and operator error.  Therefore, the
quality control component of each method is indispensable.

The data acquired from quality control procedures are used to estimate and
evaluate the information content of analytical results and to determine the
necessity for or the effect of corrective action procedures.  The parameters
used  to estimate information content include precision, accuracy, detection
limit, and other quantitative and qualitative indicators.  In addition, it
gives an overview of the activities required in an integrated program to
generate data of known and documented quality required to meet defined
objectives.

The necessary components of a complete QA/QC program include internal QC
criteria that demonstrate acceptable levels of performance, as determined by
QA review.  External review of data and procedures is accomplished by the
monitoring activities of the National Program Office, Regional data users,
Sample Management Office, NEIC, and EMSL/LV.  Each external review
accomplishes a different purpose.  These reviews are described in specific
sections of this Exhibit.  Laboratory evaluation samples, magnetic tape
audits, and data package audits provide an external QA reference for  the
program.  A laboratory on-site evaluation system is also part of the  external
QA monitoring.  A feedback loop provides the results of the various review
functions to the contract laboratories through direct communications with the
Technical Project Officers and Administrative Project Officers.

This Exhibit is not a guide to constructing quality assurance project plans,
quality control systems, or a quality assurance organization.  It is,
however, an explanation of the quality control and quality assurance
requirements of the program.  It outlines some minimum standards for QA/QC
programs.  It also includes specific items that are required in a QA Plan and
by the QA/QC documentation detailed in this contract.  Delivery of this
documentation provides the Agency with a complete data package which will
stand alone, and limits the need for contact with the Contractor or with an
analyst, at a later date, if some aspect of the analysis is questioned.

In order to assure that the product delivered by the Contractor meets the
requirements of the contract, and to improve interlaboratory data comparison,
the Agency requires the following from the Contractor:
                                     E-4                               OLM01.0

-------
A written Quality Assurance Plan, the elements of which are designated in
Section II.

Written preparation of and adherence to QA/QC Standard Operating
Procedures (SOPs) as described in Section III.
Adherence to the analytical methods and associated QC requirements
specified in the contract.

Verification of analytical standard and documentation of the purity of
neat materials and the purity and accuracy of solutions obtained from
private chemical supply houses.

Submission of all raw data and pertinent documentation for Regional
review.

Participation in the analysis of Laboratory Evaluation Samples, including
adherence to corrective action procedures.

Submission, upon request, of GC/MS tapes and applicable documentation for
tape audits.

Participation in On-Site Laboratory Evaluations,  including adherence to
corrective action procedures.

Submission of all original documentation generated during sample analyses
for Agency review.
                                E-5                         OLM01.2  1/91

-------
                                  SECTION II


                            QUALITY ASSURANCE FLAN


The Contractor shall establish a quality assurance program with the objective
of providing sound analytical chemical measurements.  This program shall
incorporate the quality control procedures, any necessary corrective action,
and all documentation required during data collection as well as the quality
assessment measures performed by management to ensure acceptable data
production.

As evidence of such a program, the Contractor shall prepare a written Quality
Assurance Plan (QAP) which describes the procedures that are implemented to
achieve the following:

o   Maintain data integrity,  validity,  and useability.

o   Ensure that analytical measurement  systems  are maintained in an
    acceptable state of stability and reproducibility .
o   Detect problems through data assessment and establishes  corrective  action
    procedures which keep  the analytical process reliable.

o   Document all  aspects of the  measurement process in order to  provide data
    which are technically  sound  and legally defensible.
The QAP must present, in specific terms, the policies, organization,
objectives, functional guidelines, and specific QA and QC activities designed
to achieve the data quality requirements in this contract.  Where applicable,
SOPs pertaining to each element shall be included or referenced as part of
the QAP.  The QAP must be available during On-Site Laboratory evaluation.
Additional information relevant to the preparation of a QAP can be found in
EPA and ASTM publications.

Elements of a Quality Assurance Plan

A.    Organization and Personnel

      1.    QA Policy and Objectives

      2.    QA Management

            a.    Organization

            b.    Assignment of QC and QA Responsibilities

            c.    Reporting Relationships

            d.    QA Document Control Procedures

            e.    QA Program Assessment Procedures

      3.    Personnel

            a.    Resumes

            b.    Education and Experience Pertinent to This Contract

            c.    Training Progress


                                    E-6                         OLM01.2  1/91

-------
Facilities and Equipment
1.    Instrumentation and Backup Alternatives
2.    Maintenance Activities and Schedules
Document Control
1.    Laboratory Notebook Policy
2.    Samples Tracking/Custody Procedures
3.    Logbook Maintenance and Archiving Procedures
4.    Case File Organization, Preparation and Review Procedures.
5.    Procedures for Preparation, Approval, Review, Revision, and
      Distribution of SOPs.
6.    Process for Revision of Technical or Documentation Procedures
Analytical Methodology
1.    Calibration Procedures and Frequency
2.    Sample Preparation/Extraction Procedures
3.    Sample Analysis Procedures
4.    Standards Preparation Procedures
5.    Decision Processes, Procedures, and Responsibility for Initiation
      of Corrective Action
Data Generation
1.    Data Collection Procedures
2.    Data Reduction Procedures
3.    Data Validation Procedures
4.    Data Reporting and Authorization Procedures
Quality Control
1.    Solvent, Reagent and Adsorbent Check Analysis
2.    Reference Material Analysis
3.    Internal Quality Control Checks
4.    Corrective Action and Determination of QC Limit Procedures
5.    Responsibility Designation
Quality Assurance
1.    Data Quality Assurance
2.    Systems/Internal Audits
3.    Performance/External Audits
4.    Corrective Action Procedures
5.    Quality Assurance Reporting Procedures
6.    Responsibility Designation


                             E-7                                OLM01.0

-------
Updating and Submission of the QAP:

Within 60 Days of contract award:

During the contract solicitation process, the Contractor was required to
submit their QAP to EMSL/LV and NEIC.  Within sixty (60) days after contract
award, the Contractor shall send a revised QAP, fully compliant with the
requirements of this contract, to the Technical Project Officer, EMSL/LV and
NEIC.  The revised QAP will become the official QAP under the contract.  The
revised QAP must include:

1)   Changes resulting from A)  The Contractor's internal review of their
     organization, personnel, facility,  equipment, policy and procedures and
     B)  The Contractor's implementation of the requirements of the contract;
     and,

2)   Changes resulting from the Agency's review of the laboratory evaluation
     sample data,  bidder supplied documentation, and recommendations made
     during the pre-award On-Site laboratory evaluation

Subsequent submissions:

During the term of contract, the Contractor shall amend the QAP when the
following circumstances occur:

1)   The Agency modifies the contract,

2)   The Agency notifies the Contractor of deficiencies in the QAP document

3)   The Agency notifies the Contractor of deficiencies resulting from the
     Agency's review of the Contractor's performance,

4)   The Contractor identifies deficiencies resulting from their internal
     review of their QAP document,

5)   The Contractor's organization, personnel,  facility, equipment, policy or
     procedures change,

6)   The Contractor identifies deficiencies resulting from the internal
     review of their organization,  personnel,  facility, equipment, policy or
     procedures changes.

The Contractor shall amend the QAP within 30 days of when the circumstances
listed above result in a discrepancy between what was previously described in
the QAP and what is presently occurring at the Contractor's facility.

When the QAP is amended, all changes  in the QAP must be clearly marked  (e.g.,
a bar in the margin indicating where  the change is found in the document, or
highlighting the change by underlining the change, bold printing the change,
or using a different print font).  The amended section pages must have  the
date on which the changes were implemented.  The Contractor shall incorporate
all amendments to the current QAP document.  The Contractor shall archive all
amendments to the QAP document for future reference by the Agency.
                                    E-8                          OLM01.2  1/91

-------
The Contractor shall send a copy of the current QAP document within 14 days
of a request by the Technical Project Officer or Administrative Project
Officers to the designated recipients.

Corrective Action:

If a Contractor fails to adhere to the requirements listed in Section II, a
Contractor may expect, but the Agency is not limited to the following
actions:  reduction of numbers of samples sent under this contract,
suspension of sample shipment to the Contractor,  GC/MS tape audit, data
package audit, an On-Site laboratory evaluation ,  remedial laboratory
evaluation sample, and/or contract sanctions,  such as a Cure Notice.
                                   E-9                         OLK01.2  1/91

-------
                                  SECTION  III
                         STANDARD OPERATING PROCEDURES
In order to obtain reliable results, adherence to prescribed analytical
methodology is imperative. In any operation that is performed on a repetitive
basis, reproducibility is best accomplished through the use of Standard
Operating Procedures (SOPs).  As defined by the EPA, an SOP is a written
document which provides directions for the step-by-step execution of an
operation, analysis, or action which is commonly accepted as the method for
performing certain routine or repetitive tasks.

SOPs prepared by the Contractor must be functional: i.e., clear,
comprehensive, up-to-date, and sufficiently detailed to permit duplication of
results by qualified analysts.  All SOPs, as presented to the Agency, must
reflect activities as they are currently performed in the laboratory.  In
addition, all SOPs must be:

o   Consistent with current EPA regulations,  guidelines,  and the CLP
    contract's requirements.

o   Consistent with instruments manufacturers's specific  instruction  manuals.

o   Available  to  the EPA during an On-Site  Laboratory Evaluation.   A  complete
    set of SOPs shall be bound together and available for inspection  at such
    evaluations.   During On-Site Laboratory evaluations,  laboratory personnel
    may be asked  to demonstrate the application of the SOPs.

o   Capable of providing for  the development of documentation that is
    sufficiently  complete to  record the performance of all tasks required by
    the protocol.

o   Capable of demonstrating  the validity of data reported by the Contractor
    and explain the cause of  missing or inconsistent results.

o   Capable of describing the corrective measures and feedback mechanism
    utilized when analytical  results do not meet protocol requirements.

o   Reviewed regularly  and updated as necessary when contract, facility,  or
    Contractor procedural modifications are made.

o   Archived for  future reference in usability or evidentiary situations.

o   Available  at  specific work stations as  appropriate

o   Subject to a  document control procedure which precludes the  use of
    outdated or inappropriate SOPs.
                                   E-10                         OLM01.2  1/91

-------
SOP FORMAT:
The format for SOPs may vary depending upon the kind of activity for which
they are prepared, however, at a minimum, the following sections must be
included:
o   Title Page
o   Scope and Application
o   Definitions
o   Procedures
o   QC  Limits
o   Corrective Action Procedures,  Including Procedures  for  Secondary Review of
    Information  Being Generated
o   Documentation Description and  Example Forms
o   Miscellaneous Notes and Precautions
o   References

SOPS REQUIRED:
The following SOPs are required by the Agency:
1.    Evidentiary SOP
      Evidentiary SOPs for required chain-of-custody and document control are
      discussed in Exhibit F, "Specification for Written Standard Operating
      Procedures"
2.    Sample Receipt and Storage
      a.    Sample receipt and identification logbooks
      b.    Refrigerator temperature logbooks
      c.    Extract storage logbooks
      d.    Security precautions
3.    Sample preparation
      a.    Reagent purity check procedures and documentation
      b.    Extraction procedures
      c.    Extraction bench sheets
      d.    Extraction logbook maintenance
4.    Glassware  cleaning
5.    Calibration (Balances, GPC)
      a.    Procedures
      b.    Frequency requirements
      c.    Preventative maintenance schedule and procedures
      d.    Acceptance criteria and corrective actions
                                   E-ll                         OLM01.2  1/91

-------
      e.    Logbook maintenance authorization
6.    Analytical procedures (for each analytical system)
      a.    Instrument performance specifications
      b.    Instrument operating procedures
      c.    Data acquisition system operation
      d.    Procedures when automatic quantitation algorithms are overridden
      e.    QC required parameters
      f.    Analytical run/injection logbooks
      g.    Instrument error and editing flag descriptions and resulting
            corrective actions
7.    Maintenance activities (for each analytical system)
      a.    Preventative maintenance schedule and procedures
      b.    Corrective maintenance determinants and procedures
      c.    Maintenance authorization
8.    Analytical standards
      a.    Standard coding/identification and inventory system
      b.    Standards preparation logbook(s)
      c.    Standard preparation procedures
      d.    Procedures for equivalency/traceability analyses and
            documentation
      e.    Purity logbook (primary standards and solvents)
      f.    Storage, replacement, and labelling requirements
      g.    QC and corrective action measures
9.    Data reduction procedures
      a.    Data processing systems operation
      b.    Outlier identification methods
      c.    Identification of data requiring corrective action
      d.    Procedures for format and/or forms for each operation
10.   Documentation policy/procedures
      a.    Laboratory/analyst's notebook policy, including review policy
      b.    Complete SDG File contents
      c.    Complete SDG File organization and assembly procedures, including
            review policy
      d.    Document inventory procedures, including review policy
11.   Data validation/self inspection procedures
      a.    Data flow and chain-of-command for data review
      b.    Procedures for measuring precision and accuracy

                                   E-12                         OLM01.2  1/91

-------
      c.    Evaluation parameters for identifying systematic errors
      d.    Procedures to assure that hardcopy and diskette deliverables are
            complete and compliant with the requirements in SOW Exhibits B
            and H.

      e.    Procedures to assure that hardcopy deliverables are in agreement
            with their comparable diskette deliverables.

      f.    Demonstration of internal QA inspection procedure (demonstrated
            by supervisory sign-off on personal notebooks, internal
            laboratory evaluation samples, etc.).

      g.    Frequency and type of internal audits (eg.,  random, quarterly,
            spot checks, perceived trouble areas).

      h.    Demonstration of problem identification-corrective actions and
            resumption of analytical processing.  Sequence resulting from
            internal audit (i.e., QA feedback).

      i.    Documentation of audit reports, (internal and external),
            response, corrective action, etc.

12.   Data management and handling

      a.    Procedures for controlling and estimating data entry errors.

      b.    Procedures for reviewing changes to data and deliverables and
            ensuring traceability of updates.

      c.    Lifecycle management procedures for testing, modifying and
            implementing changes to existing computing systems including
            hardware, software, and documentation or installing new systems.

      d.    Database security, backup and archival procedures including
            recovery from system failures.

      e.    System maintenance procedures and response time.

      f.    Individuals(s) responsible for system operation, maintenance,
            data integrity and security.

      g.    Specifications for staff training procedures.


SOPS DELIVERY REQUIREMENTS:

Updating and submission of SOPs:

Within 60 days of contract award:

During the contract solicitation process, the Contractor was required to
submit their SOPs to EMSL/LV and NEIC.  Within sixty (60) days after contract
award, the Contractor shall send a complete revised set of SOPs, fully
compliant with the requirements of this contract, to the Technical Project
Officer, EMSL/LV and NEIC.  The revised SOPs will become the official SOPs
under the contract.  The revised SOPs must include:

1)   Changes resulting from A)  the Contractor's internal review of their
     procedures and B)  the Contractor's implementation  of the requirements
     of the contract;


                                   E-13                         OLM01.2   1/91

-------
2)   Changes resulting from the Agency's review of the laboratory evaluation
     sample data, bidder supplied documentation,  and recommendations made
     during the pre-award On-Site laboratory evaluation.

Subsequent Submissions:

During the term of contract, the Contractor shall amend the SOPs when the
following circumstances occur:

1)   The Agency modifies the contract,

2)   The Agency notifies the Contractor of deficiencies in their SOPs
     documentation

3)   The Agency notifies the Contractor of deficiencies resulting from the
     Agency's review of the Contractor's performance,

4)   The Contractor's procedures change,

5)   The Contractor identifies deficiencies resulting from the internal
     review of their SOPs documentation, or

6)   The Contractor identifies deficiencies resulting form the internal
     review of their procedures.

The SOPs must be amended or new SOPs must be written within 30 days of when
the circumstances listed above result in a discrepancy between what was
previously described in the SOPs and what is presently occurring at the
Contractor's facility.  All changes in the SOPs must be clearly marked (e.g.,
a bar in the margin indicating where the change is in the document, or
highlighting the change by underlining the change, bold printing the change,
or using a different print font).  The amended/new SOPs must have the date on
which the changes were implemented.

When the SOPs are amended or new SOPs are written, the Contractor shall
document in a letter the reasons for the changes, and submit the amended SOPs
or new SOPs to the Technical Project Officer, EMSL/LV (quality
assurance/technical SOPs) and NEIC (evidentiary SOPs).  The Contractor shall
send the letter and the amended sections of the SOPs or new SOPs within 14
days of the change.  An alternate delivery schedule for the submittal of the
letter and amended/new SOPs may be proposed by the Contractor, but it is the
sole decision of the Agency, represented either by the Technical Project
Officer or Administrative Project Officer, to approve or disapprove the
alternate delivery schedule.  If an alternate delivery schedule is proposed,
the Contractor shall describe in a letter to the Technical Project Officer,
Administrative Project Officer, and the Contracting Officer why he/she is
unable to meet the delivery schedule listed in this section.  The Technical
Project Officer/Administrative Project Officer will not grant an extension
for greater than 30 days for amending/writing new SOPs.  The Technical
Project Officer/Administrative Project Officer will not grant an extension
for greater than 14 days for submission of the letter documenting the reasons
for the changes and for submitting amended/new SOPs.  The Contractor shall
proceed and not assume that an extension will be granted until so notified by
the TPO and/or APO.

                                   E-14                         OLM01.2  1/91

-------
The Contractor shall send a complete set of current SOPs within 14 days of a
request by the Technical Project Officer or Administrative Project Officer to
the recipients he/she designates.

Corrective action:

If a Contractor fails to adhere to the requirements listed in Exhibit E,
Section III, a Contractor may expect, but the Agency is not limited to the
following action:  reduction of number of samples sent under this contract,
suspension of sample shipment to the Contractor, GC/MS tape audit, data
package audit, On-Site laboratory evaluation, remedial laboratory evaluation
sample, and/or contract sanction, such as a Cure Notice.
                                   E-15                        OLM01.2  1/91

-------
                          SECTION IV VGA
                    VOLATILE QA/QC REQUIREMENTS
                           INTRODUCTION

Sections II and III of this exhibit outline the requirements for the
quality assurance program that each laboratory must establish under
this contract.  The purpose of Section III is to outline the minimum
quality control (QC) operations necessary to satisfy the analytical
requirements associated with the determination of the volatile organic
target compounds listed in Exhibit C using the procedures in Exhibit D
VOA for water and soil/sediment samples.  This section is not intended
as a comprehensive quality control document, but rather as a guide to
the specific QC operations that must be considered for volatile
analyses.  At a minimum, the laboratory is expected to address these
operations in preparing the quality assurance plan and QA/QC Standard
Operating Procedures discussed in Section II.

These operations include the following:

o   GC/MS Mass Calibration and Ion Abundance Patterns
o   GC/MS Initial  and Continuing Calibration
o   Stability of Internal  Standard Responses and Retention Times
o   Method Blank Analysis
o   System Monitoring Compound Recoveries
o   Matrix Spike and Matrix Spike Duplicate  Analyses
o   Dilution of Samples, Matrix Spikes,  and  Matrix Spike  Duplicates
Not discussed in this section are the requirements for quality
assurance of the data reporting aspects of volatile analyses which are
described in general terms in Section II and III of this exhibit.

GC/MS Mass Calibration and Ion Abundance Patterns

Prior to initiating any data collection activities involving samples,
blanks, or standards, it is necessary to establish that a given GC/MS
system meets the instrument performance criteria specified in Exhibit D
VOA, Section IV, paragraph 6.  The purpose of this instrument
performance check is to assure correct mass calibration, mass
resolution, and mass transmission.  This is accomplished through the
analysis of Bromofluorobenzene (BFB).

1.1     The required frequency of BFB  analysis  (once  every 12 hours  on
        each GC/MS  system)  is  described in  detail in  Ex.  D VOA,  Section
        IV, paragraph 6.4.

1.2     The key ions  produced  during the analysis of  BFB and their
        respective  ion abundance  criteria are given in Table 1,  Ex.  D
        VOA, Section IV, paragraph 6.4.4.
                           E-16/VOA                             OLM01.0

-------
      1.3     The documentation includes Form V VGA, and a mass listing and
              bar graph spectrum of each BFB analysis.

2.    GC/MS Initial Calibration for Target Compounds and System Monitoring
      Compounds

      Prior to the analysis of samples and required blanks and after
      instrument performance criteria have been met, the GC/MS system must be
      initially calibrated at a minimum of five concentrations to determine
      the linearity of response utilizing target compound and system
      monitoring compound standards.

      2.1     The concentrations of the initial calibration standards for
              volatile target compounds and system  monitoring compounds are
              10, 20, 50,  100, and 200 ug/L, as described in Ex. D VGA,
              Section IV,  paragraph 5.5.

      2.2     The standards are to be analyzed according to the procedures
              given in  Ex. D VOA, Section IV, paragraph 7, and at the
              frequency given in that paragraph.

      2.3     The relative response factors (RRFs)  are determined according
              to the procedures in Ex. D VOA, Section IV, paragraph 7.4,
              using the assignment of internal standard to target compounds
              and system monitoring compounds given in Ex. D VOA, Section  IV,
              paragraph 7.4, and Table 5.

      2.4     The calibration of the GC/MS is evaluated on the basis of the
              magnitude and stability of the relative response factors of
              each target  compound and system monitoring compound.  The
              minimum RRF  of each compound at each  concentration level in  the
              initial calibration and the percent relative standard deviation
              (%RSD) across all five points must meet the criteria given Ex.
              D VOA, Section IV, paragraphs 7.4.5 and 7.4.6, and Table 2.
              Allowance is made for any two volatile compounds that fail to
              meet these criteria.  The minimum RRFs of those two compounds
              must be greater than or equal to 0.010, and the %RSD must be
              less than or equal to 40.0% for the initial calibration to be
              acceptable.

      2.5     The documentation includes Form VI VOA, a GC/MS data system
              printout for the analysis of each volatile calibration
              standard.

3.    GC/MS Continuing Calibration for Target Compounds  and System Monitoring
      Compounds.

      Once the GC/MS system has been calibrated,  the calibration must be
      verified each twelve (12)  hour time period for each GC/MS system.

      3.1     The concentration of the continuing calibration standard for
              volatile target compounds and system  monitoring compounds is 50
              ug/L, as described in Ex. D VOA, Section IV, paragraph 5.5.3.


                                 E-17/VOA                       OLM01.4  3/91

-------
      3.2     The standard is to be analyzed according to the procedures
              given  in  Ex. D VGA, Section IV, paragraph 7, and at the
              frequency given in that paragraph.

      3.3     The continuing calibration of the GC/MS system is evaluated on
              the basis of the magnitude of the relative response factors and
              the percent difference between the'average RRF of each compound
              from the initial calibration and the RRF of that compound in
              the continuing calibration standard.  The minimum RRF of each
              compound in the continuing calibration and the percent
              difference must meet the criteria given Ex. D VOA, Section IV,
              paragraphs 7.4.5, 7.4.6 and 7.4.7, and Table 2.  Allowance is
              made for any two volatile compounds that fail to meet these
              criteria.  The minimum RRFs of those two compounds must be
              greater than or equal to 0.010, and the percent difference must
              be less than or equal to 40.0% for the continuing calibration
              to be  acceptable.

      3.4     The documentation includes Form VII VOA, a GC/MS data system
              printout for the analysis of the volatile calibration standard.

4.    Internal Standard Responses and Retention Times

      The response of each of the internal standards in all calibration
      standards, samples,  and blanks is crucial to the provision of reliable
      analytical results,  because the quantitative determination of volatile
      compounds by these procedures is based on the use of internal standards
      added immediately prior to analysis.

      4.1     The specific compounds used as internal standards are given in
              Ex. D  VOA, Section IV, paragraph 5.4.3.  The concentration of
              each internal standard in the aliquot of the sample analyzed by
              GC/MS  must be 50 ug/L at the time of purging.

      4.2     The retention time and the extracted ion current profile  (EICP)
              of each internal standard must be monitored  for all analyses.

      4.3     The area response of each internal standard  from the EICP and
              the retention time of the internal standard  are evaluated for
              stability, according to the procedures in Ex. D VOA, Section
              IV, paragraph 10.2.  The area of the internal standard  in a
              sample must not vary by more than a factor of 2 (i.e. -50% to
              +100%) from the area of the same internal standard in the
              associated continuing calibration standard.  Likewise,  the
              retention time of an internal standard must  be within +0.50
              minutes (30 seconds) of its retention  time  in the continuing
              calibration standard  (see Ex.D VOA, Section  IV, paragraph
              10.2).

      4.4     Requirements for reanalysis of samples when  internal standards
              do not meet specifications are given  in Ex.  D VOA, Section IV,
              paragraph 10.2.
                                 E-18/VOA                       OLM01.4  3/91

-------
      4.5      The documentation  includes Form VIII VOA, and the GC/MS data
               system printout  for the analysis of each sample, blank, matrix
               spike,  matrix  spike duplicate, and standard.

5.    Method Blank Analysis

      A method blank is a volume of a clean reference matrix (deionized
      distilled water for water samples, or a purified solid matrix for
      soil/sediment samples) that is carried through the entire analytical
      procedure.  The volume or weight of the reference matrix must be
      approximately equal to the volume or weight of samples associated with
      the blank.  The purpose of a method blank is to determine the levels of
      contamination associated with the processing and analysis of samples.

      5.1      For volatile analysis, a method blank must be analyzed once
               every  12 hours on  each GC/MS system, as described in detail in
               Ex.  D  VOA, Section IV, paragraphs 8.1.17, 8.2.1.9, and
               8.2.2.10.

      5.2      For the purposes of this protocol, an acceptable method blank
               must meet the  criteria in paragraphs 5.2.1. and 5.2.2 below.

               5.2.1      A method blank for volatile analysis must contain
                         less than or equal to five times (5x) the Contract
                         Required Quantitation Limit (CRQL, see Exhibit C) of
                         Methylene chloride, Acetone, and 2-Butanone.

               5.2.2      For all other target compounds, the method blank
                         must contain less than or equal to the Contract
                         Required Quantitation Limit (CRQL, see Exhibit C) of
                         any single target compound.

      5.3      If  a method blank  exceeds the limits for contamination above,
               the  Contractor must consider the analytical system out of
               control.  The  source of the contamination must be investigated
               and appropriate corrective actions taken and documented before
               further sample analysis proceeds.  The requirements for
               reanalysis of  associated samples are given in Ex. D VOA,
               Section IV, paragraph 10.10.

      5.4      The  documentation  includes Form I VOA for the blank analysis,
               Form IV VOA, associating the samples and the blank, and a GC/MS
               data system printout for the analysis of the method blank.

6.     System Monitoring Compound Recoveries

      The recoveries  of the three system monitoring compounds are calculated
      from the analysis of each sample,  blank,  matrix spike and matrix spike
      duplicate.  The purpose of the system monitoring compounds is to
      evaluate the performance of the entire purge and trap-gas
      chromatograph-mass spectrometer system.   Poor purging efficiency,
      leaks,  and cold spots in transfer lines are only a few of the potential
      causes  of poor  recovery of these compounds.
                                 E-19/VOA                      OLM01.2   1/91

-------
      6.1      The system monitoring compounds are added  to each sample,
               blank,  matrix spike,  and matrix spike  duplicate prior to
               purging or extraction (medium  soils only), at  the concentration
               described in Ex.  D VGA,  Section IV, paragraph  5.4.4.

      6.2      The recoveries of the system monitoring compounds are
               calculated according  to  the procedures in  Ex.  D VOA, Section
               IV, paragraph 10.8.1.

      6.3      The recoveries must be within  the quality  control limits  given
               in Ex.  D VOA,.  Section IV, Table 6.  If the recovery of any one
               system monitoring compound is  outside  these limits, the
               Contractor must follow the steps outlined  in Ex. D VOA, Section
               IV, paragraphs 10.8.2 to 10.8.6.

      6.4      The documentation includes Form II VOA, and a  GC/MS data  system
               printout for the analysis of each sample,  blank, matrix spike,
               and matrix spike duplicate.

7.    Matrix Spike and Matrix Spike Duplicate Analysis

      In order to evaluate the effects of the sample matrix on the methods
      used for volatile analyses,  the Agency has prescribed a mixture of
      volatile target compounds to be spiked into two aliquots of a sample,
      and analyzed in accordance with the appropriate method.

      7.1      The frequency of matrix  spike  and matrix spike duplicate
               (MS/MSD)  analysis is  described in Ex.  D VOA, Section IV,
               paragraph 10.9.

      7.2      The recoveries of the matrix spike compounds are calculated
               according to the procedures in Ex. D VOA,  Section IV, paragraph
               10.9.1.   The relative percent  difference between the results
               for each spiked analyte  of the matrix  spike and the matrix
               spike  duplicate are calculated according to the procedures in
               Ex.  D  VOA,  Section IV, paragraph 10.9.2.

      7.3      The quality control limits for recovery and relative percent
               difference  are given  in  Ex. D  VOA, Section IV, Table 7.   These
               limits  are  only advisory at this time, and no  further action  is
               required when the limits are exceeded.

      7.4      The documentation includes Form I VOA  for  both the MS and MSD
               analyses,  Form III VOA,  and a  GC/MS printout for each analysis.

8.    Dilution of Samples, Matrix Spikes,  and Matrix Spike Duplicates

      If the on-column concentration of any sample exceeds the initial
      calibration range, that sample must be diluted and reanalyzed,  as
      described in Ex. D VOA, Section IV,  paragraph 10.7.  Guidance in
      performing dilutions and exceptions are given in that paragraph, and
      reiterated here.
                                 E-20/VOA                       OLM01.2  1/91

-------
8.1      Use  the  results  of the  original analysis  to determine  the
         approximate  dilution factor  required  to get the largest  analyte
         peak within  the  initial calibration range.

8.2      The  dilution factor chosen should keep the response of the
         largest  analyte  peak for a target compound in  the upper  half of
         the  initial  calibration range of the  instrument.

8.3      Do not submit data for  more  than two  analyses, i.e., the
         original sample  and one dilution, or, if  the volatile  screening
         procedure was employed,  from the most concentrated dilution
         analyzed and one further dilution.

8.4      Do not dilute MS/MSD samples to get either spiked or non-spiked
         analytes within  calibration  range. If the sample from  which the
         MS/MSD aliquots  were taken contains high  levels of the spiked
         analytes, calculate the concentration and recovery of  the
         analytes from the undiluted  analysis, and note the problem in
         the  SDG  Narrative.

8.5.      For  total Xylenes,  where three isomers are quantified  as two
         peaks, the calibration  of each peak,  should be considered
         separately,  i.e.,  a diluted  analysis  is not required for total
         Xylenes  unless the concentration of either peak separately
         exceeds  200  ug/L.
                           E-21/VOA                     OLM01.1   12/90

-------
                                SECTION IV SV
                       SEMIVOLATILE QA/QC REQUIREMENTS
                                 INTRODUCTION

      Sections II and III of this exhibit outline the requirements for the
      quality assurance program that each laboratory must establish under
      this contract.  The purpose of Section III is to outline the minimum
      quality'control (QC) operations necessary to satisfy the analytical
      requirements associated with the determination of the semivolatile
      organic target compounds listed in Exhibit C using the procedures in
      Exhibit D SV for water and soil/sediment samples.   This section is not
      intended as a comprehensive quality control document, but rather as a
      guide to the specific QC operations that must be considered for
      semivolatile analyses.  At a minimum, the laboratory is expected to
      address these operations in preparing the quality assurance plan and
      QA/QC Standard Operating Procedures discussed in Section II.

      These operations include the following:

      o   GC/MS Mass Calibration and Ion Abundance  Patterns
      o   GC/MS Initial  and Continuing Calibration
      o   Stability of Internal  Standard Responses  and Retention Times

      o   Method Blank Analysis
      o   Surrogate Recoveries
      o   Matrix Spike and Matrix Spike Duplicate Analyses
      o   Dilution of Samples, Matrix Spikes,  and Matrix Spike Duplicates

      Not discussed in this section are the requirements for quality
      assurance of the data reporting aspects of semivolatile analyses which
      are described in general terms in Section II and III of this exhibit.

1.     GC/MS Mass Calibration and Ion Abundance Patterns

      Prior to initiating any data collection activities involving samples,
      blanks,  or standards, it is necessary to establish that a given GC/MS
      system meets the instrument performance criteria specified in Exhibit D
      SV, Section IV, paragraph 4.  The purpose of this instrument
      performance check is to assure correct mass calibration, mass
      resolution, and mass transmission.  This is accomplished through the
      analysis of Decafluorotriphenyl phosphine (DFTPP).

      1.1     The  required  frequency  of  DFTPP analysis  (once  every  12 hours
              on each GC/MS system)  is described in detail in Ex. D SV,
              Section IV, paragraph 4.3.6.

      1.2     The key ions produced during  the analysis  of DFTPP and their
              respective  ion abundance criteria  are given  in  Table  1, Ex. D
              SV,  Section IV, paragraph  4.3.3.
                                 E-22/SV                              OLM01.0

-------
      1.3     The documentation includes Form V SV, and a mass listing and
              bar graph spectrum of each DFTPP analysis.

2.    GC/MS Initial Calibration for Target Compounds and Surrogates.

      Prior to the analysis of samples and required blanks,  and after
      instrument performance criteria have been met, the GC/MS system must be
      initially calibrated at a minimum of five concentrations to determine
      the linearity of response utilizing target compound and surrogate
      standards.

      2.1     The levels of  the initial calibration standards for
              semivolatile target compounds and surrogates are  20, 50, 80,
              120, and 160 ng, in a 2 uL injection volume, as described in
              Ex. D SV, Section IV, paragraph 3.2.

      2.2     The standards  are to be analyzed according to the procedures
              given in  Ex.  D SV, Section IV, paragraph 5, and at the
              frequency given in that paragraph.

      2.3     The relative response factors (RRFs)  are determined according
              to the procedures in Ex. D SV, Section IV, paragraph 5.4, using
              the assignment of internal standard to target compounds and
              surrogates given in Ex. D SV, Section IV, paragraph 5.4, and
              Tables 3 and 4.

      2.4     The calibration of the GC/MS is evaluated on the basis of the
              magnitude and  stability of the relative response factors of
              each target compound and surrogate.  The minimum RRF of each
              compound at each concentration level  in the initial calibration
              and the percent relative standard deviation (%RSD) across all
              five points must meet the criteria given Ex. D SV, Section IV,
              paragraph 5.6, and Table 5.  Allowance is made for any four
              semivolatile compounds that fail to meet these criteria.  The
              minimum PxRFs of those four compounds must be greater than or
              equal to 0.010, and the %RSD must be  less than or equal to
              40.0% for the  initial calibration to be acceptable.

      2.5     The documentation includes Form VI SV, a GC/MS data system
              printout for the analysis of each semivolatile calibration
              standard.

3.    GC/MS Continuing Calibration for Target Compounds and Surrogates.

      Once the GC/MS system has been calibrated,  the calibration must be
      verified each twelve (12)  hour time period for each GC/MS system.

      3.1     The level of the continuing calibration standard for
              semivolatile target compounds and surrogates is 50 ng, in a 2
              uL injection volume, as described in  Ex. D SV, Section IV,
              paragraph 3.2.
                                  E-23/SV                       OLM01.4   3/91

-------
      3.2     The standard is to be analyzed according to the procedures
              given in  Ex. D SV, Section IV, paragraph 5, and at the
              frequency given in that paragraph.

      3.3     The continuing calibration of the GC/MS system is evaluated on
              the basis of the magnitude of the relative response factors and
              the percent difference between the average RRF of each compound
              from the initial calibration and the RRF of that compound in
              the continuing calibration standard.  The minimum RRF of each
              compound in the continuing calibration and the percent
              difference must meet the criteria given Ex. D SV, Section IV,
              paragraphs 5.6, 5.7, and Table 5.  Allowance is made for any
              four semivolatile compounds that fail to meet these criteria.
              The minimum RRFs of those four compounds must be greater than
              or equal to 0.010, and the %D must be less than or equal to
              40.0% for the continuing calibration to be acceptable.

      3.4     The documentation includes Form VII SV, a GC/MS data system
              printout for the analysis of the semivolatile calibration
              standard.

4.    Internal Standard Responses and Retention Times

      The response of each of the internal standards in all calibration
      standards, samples,  and blanks is crucial to the provision of reliable
      analytical results  because the quantitative determination of
      semivolatile compounds by these procedures is based on the use of
      internal standards  added immediately prior to analysis.

      4.1     The specific compounds used as internal standards are given in
              Ex. D SV, Section IV, paragraph 3.1.  The amount of each
              internal standard in the injection volume (2 uL) of the sample
              extract analyzed by GC/MS must be 40 ng (20 ng/uL).

      4.2     The retention time and the extracted ion current profile (EICP)
              of each internal standard must be monitored for all analyses.

      4.3     The area response of each internal standard from the EICP and
              the retention time of the internal standard are evaluated for
              stability, according to the procedures  in Ex. D SV, Section IV,
              paragraph 8.1.  The area of the internal standard in a sample
              must not vary by more than a factor of  2 (i.e.  -50% to +100%)
              from the area of the same internal standard in  the associated
              continuing calibration standard.  Likewise, the retention time
              of an internal standard must be within  +0.50 minutes  (30
              seconds) of its retention time in the continuing calibration
              standard (see Ex.D SV, Section IV, paragraph 8.1).

      4.4     Requirements for reanalysis of samples  when internal standards
              do not meet specifications are given in Ex. D SV, Section IV,
              paragraph 8.1.
                                  E-24/SV                       OLM01.4   3/91

-------
      4.5     The documentation includes Form VIII SV, and the GC/MS data
              system printout for the analysis of each sample, blank, matrix
              spike, matrix spike duplicate, and standard.

5.    Method Blank Analysis

      A method blank is a volume of a clean reference matrix (deionized
      distilled water for water samples, or a purified sodium sulfate for
      soil/sediment samples)  that is carried through the entire analytical
      procedure.  The volume or weight of the reference matrix must be
      approximately equal to the volume or weight of samples associated with
      the blank.  The purpose of a method blank is to determine the levels of
      contamination associated with the processing and analysis of samples.

      5.1     For semivolatile analysis, one method blank must be extracted
              with each group of samples of a similar matrix and
              concentration level (soils only), as described in Ex. D SV,
              Section IV, paragraph 8.7.

      5.2     For the purposes of this protocol, an acceptable method blank
              must meet the criteria in paragraphs 5.2.1. and 5.2.2 below.

              5.2.1      A method blank for semivolatile analysis must
                         contain less than or equal to five times (5x) the
                         Contract Required Quantitation Limit (CRQL, see Ex.
                         C) of the phthalate esters listed in Ex. C.

              5.2.2      For all other target compounds, the method blank
                         must contain less than or equal to the Contract
                         Required Quantitation Limit (CRQL, see Exhibit C) of
                         any single target compound.

      5.3     If a method blank exceeds the limits for contamination above,
              the Contractor must consider the analytical system out of
              control.  The source of the contamination must be  investigated
              and appropriate corrective actions taken and documented before
              further sample analysis proceeds.  The requirements for
              reextraction and reanalysis of associated samples  are given in
              Ex. D SV, Section IV, paragraph 8.7.

      5.4     The documentation includes Form I SV for'the blank analysis,
              Form IV SV, associating the samples and the blank, and a GC/MS
              data system printout for  the analysis of the method blank.

6.    Surrogate Recoveries

      The recoveries of the eight surrogates are calculated from the analysis
      of each sample,  blank,  matrix spike and matrix spike duplicate.  The
      purpose of the surrogates is to evaluate the preparation and analysis
      of samples.
                                 E-25/SV                              OLM01.0

-------
      6.1      The  surrogates  are added to each sample, blank, matrix .spike,
               and  matrix  spike duplicate prior to extraction, at the
               concentrations  described in Ex. D SV, Section II, Part B
               paragraph 4.6 and Part C paragraph 1.5.3 and 2.4.5.

      6.2      The  recoveries  of the surrogates are calculated according  to
               the  procedures  in Ex. D SV, Section IV, paragraph 8.5.1.

      6.3      The  recoveries  must be within the quality control limits given
               in Ex.  D SV, Section IV, Table 6.  If the recovery of any
               surrogate is outside these limits, the Contractor must follow
               the  steps outlined in E. D SV, Section IV, paragraphs 8.5.2 to
               8.5.6 to determine whether or not reextraction and/or
               reanalysis  is required.

      6.4      The  documentation includes Form II SV, and a GC/MS data system
               printout for the analysis of each sample, blank, matrix spike,
               and  matrix  spike duplicate.

7.    Matrix Spike and Matrix Spike Duplicate Analysis

      In order to evaluate the effects of the sample matrix on the methods
      used for semivolatile analyses,  the Agency has prescribed a mixture of
      semivolatile target compounds to be spiked into two aliquots of a
      sample, and analyzed in accordance with the appropriate method.

      7.1      The  frequency of matrix spike and matrix spike duplicate
               (MS/MSD) analysis is described in Ex. D SV, Section  IV,
               paragraph 8.6.

      7.2      The  recoveries  of the matrix spike compounds are calculated
               according to the procedures in Ex. D SV, Section IV, paragraph
               8.6.1.  The relative percent difference between the  results for
               each spiked analyte of the matrix spike and the matrix spike
               duplicate is calculated according to the procedures  in Ex. D
               SV,  Section IV, paragraph 8.6.2.

      7.3      The  quality control limits for recovery and relative percent
               difference  are  given in Ex. D SV, Section IV, Table  7.  These
               limits  are  only advisory at this time, and no further action is
               required when the limits are exceeded.

      7.4      The  documentation includes Form I SV for both the MS and MSD
               analyses, Form  III SV, and a GC/MS printout for each analysis.

8.    Dilution of Samples, Matrix Spikes,  and Matrix Spike Duplicates

      If the on-column concentration of any sample exceeds the initial
      calibration range,  that sample must be diluted and reanalyzed,  as
      described in Ex. D SV,  Section IV,  paragraph 8.4.   Guidance  in
      performing dilutions and exceptions are given in that paragraph,  and
      reiterated here.
                                  E-26/SV                        OLM01.2  1/91

-------
8.1     Use  the  results  of  the original analysis to determine  the
        approximate  dilution  factor required  to get the  largest  analyte
        peak within  the  initial calibration range.

8.2     The  dilution factor chosen should keep the response  of the
        largest  analyte  peak  for a target compound in  the upper  half  of
        the  initial  calibration range of the  instrument.

8.3     Do not submit data  for more than two  analyses, i.e., the
        original sample  and one dilution, or, if the semivolatile
        screening procedure was employed, from the most  concentrated
        dilution analyzed and one further dilution.

8.4     Do not dilute MS/MSD  samples to get either spiked or non-spiked
        analytes within  the calibration range. If the  sample from which
        the  MS/MSD aliquots were taken contains high levels  of the
        spiked analytes, calculate the concentration and recovery of
        the  analytes  from the undiluted analysis, and  note the problem
        in the SDG Narrative.
                           E-27/SV                      OLM01.1  12/90

-------
                               SECTION IV PEST
                     PESTICIDE/AROCLOR QA/QC REQUIREMENTS
                                    INTRODUCTION

      Sections II and III of this exhibit outline the requirements for the
      quality assurance program that each laboratory must establish under
      this contract.  The purpose of Section III is to outline the minimum
      quality control (QC) operations necessary to satisfy the analytical
      requirements associated with the determination of the pesticide/Aroclor
      target compounds listed in Exhibit C using the procedures in Exhibit D
      PEST for water and soil/sediment samples.  This section is not intended
      as a comprehensive quality control document, but rather as a guide to
      the specific QC operations that must be considered for
      pesticide/Aroclor analyses.  At a minimum, the laboratory is expected
      to address these operations in preparing the quality assurance plan and
      QA/QC Standard Operating Procedures discussed in Section II.

      These operations include the following:

      o   GC  Column  Resolution

      o   GC/EC  Initial  and Continuing Calibration

      o   Determination  of Retention Times  and Retention Time  Windows

      o   Analytical Sequence

      o   Blank  Analyses

      o   Surrogate  Recoveries

      o   Matrix Spike and Matrix Spike Duplicate  Analyses

      o   Dilution of Samples, Matrix Spikes,  and  Matrix Spike Duplicates

      Not discussed in this section are the requirements for quality
      assurance of the data reporting aspects of pesticide/Aroclor analyses
      which are described in general terms in Section II and III of this
      exhibit.

1.     GC Column Resolution

      Prior to initiating any data collection activities involving samples,
      blanks, or standards, it is necessary to establish that a given GC
      column meets  the analyte resolution criteria specified in Exhibit D
      PEST,  Section III, paragraph 6.2.2.  The purpose of this resolution
      check is to demonstrate that at the time of the initial calibration,
      the GC column is capable of chromatographically resolving the target
      compounds.  This is accomplished through the analysis of the Resolution
      Check Mixture (see Ex. D,  Section III, paragraph 3.1), which contains
      the nine target compounds  that are most difficult to resolve.

      1.1     The Resolution Check Mixture  must be analyzed at the  beginning
              of every  initial calibration  sequence, on each  GC  column and
              instrument  used for analysis.


                                 E-28/PEST                             OLM01.0

-------
      1.2     Additional resolution criteria apply to the target compounds in
              the  standards used for initial calibration and calibration
              verification, as described in Ex. D Section III, paragraphs
              6.2.4, 6.2.10, and 7.10.

      1.3     The  documentation includes Form VI FEST-4, chromatograms and
              data system printouts for the analysis of the Resolution Check
              Mixture on each GC column and instrument used for analysis.

2.    GC/EC Initial Calibration for Target Compounds  and Surrogates.

      Prior to the analysis of samples and required blanks,  the GC/EC system
      must be initially calibrated at a minimum of three concentrations to
      determine the linearity of response utilizing single component target
      compound and surrogate standards.  Multicomponent target compounds are
      calibrated at a single point.

      2.1     The  concentrations of the low point initial calibration
              standards for single component pesticide target compounds and
              surrogates are described in Ex. D PEST, Section III, paragraph
              3.3.  The concentration of the mid point initial calibration
              standards is specified in Ex. D, Section III, paragraph 3.3 as
              4 times the low point concentration.  The concentration of the
              high point initial calibration standard must be at least 16
              times the low point concentration, and may be higher as
              described in Ex. D PEST, Section III, paragraph 3.3.

      2.2     The  standards are to be analyzed according to the procedures
              given in  Ex. D PEST, Section III, using the GC operating
              conditions in paragraphs 4 and 6, and at the frequency given in
              paragraph 6.1.

      2.3     The  calibration factors are determined according to the
              procedures in Ex. D PEST, Section III,  paragraphs 9 and 10.

      2.4     The  initial calibration of the GC/EC is evaluated on the basis
              of the stability of the calibrations factors and retention
              times of each target compound and surrogate, described in Ex. D
              PEST, Section III, paragraphs 6.2.5 to 6.2.9.

      2.5     The  calibration is also evaluated on the basis of the extent of
              breakdown of two target compounds, Endrin and 4,4'-DDT, as
              described in Ex. D PEST, Section III, paragraph 6.2.3.

      2.6     The  documentation includes Form VI PEST, chromatograms and data
              system printouts of all standards for the pesticide/Aroclor
              calibration standards.
                                E-29/PEST                      OLM01.2   1/91

-------
 3.    GC/EC Continuing Calibration for Target compounds and Surrogates.

      Once the GC/EC system has been calibrated, the calibration must be
      verified each twelve (12) hour time period for each GC column and
      instrument used for analysis.  The calibration is verified through the
      analysis of  instrument blanks, Performance Evaluation Mixtures (PEM),
      and the mid  point concentrations of Individual Standard Mixtures A and
      B.

      3.1      The concentrations of the  PEM and Individual  Standard Mixtures
               used for continuing calibration are  given in  Ex.  D PEST,
               Section III,  paragraphs  3.2  and 3.3.

      3.2      The instrument blank is  described in Ex.  D PEST,  Section III,
               paragraph 15.3.

      3.3      The instrument blank and the  standards must be  analyzed  once
               every twelve  hours according  to the  procedures  in Ex. D  PEST,
               Section III,  paragraph 5,  bracketing the  sample analyses,  as
               described in  Ex.  0 PEST,  Section III, paragraph 7.

      3.4      The continuing calibration is evaluated on the  basis of  the
               stability of  the  retention times of  the target  compounds in  the
               standards.

      3.5      The continuing calibration is evaluated on the  basis of  the
               stability of  the  instrument response to the target compounds  in
               the PEM,  as judged by the  reproducibility of  the  determinations
               of  the  concentrations of these compounds  in the  standard,  as
               described in  Ex.  D PEST,  Section III, paragraph  7.10.

      3.6      The continuing calibration is evaluated on the basis of  the
               extent  of breakdown of two target compounds in  the PEM,  Endrin
               and 4,4'-DDT,  as  described in Ex.  D  PEST,  Section III,
               paragraph 7.10.

      3.7      The  continuing calibration is evaluated'on the basis of  the
               levels  of contamination  that  are found in the instrument blank,
               as  described in Ex.  D PEST, Section  III paragraph 15.3.

      3.8      The  documentation includes Form VII  PEST,  Form  VIII PEST,
               chromatograms  and data system printouts for all  standards  and
               instrument blanks analyzed.

4.     Determination of Retention Times and Retention Time Windows

      The identification of single component pesticides by gas
      chromatographic methods is based primarily on retention time data.  The
      identification of multicomponent analytes is based primarily on
      recognition of patterns of retention  times displayed on a chromatogram.
      Therefore,  the determination of retention times and retention time
      windows is crucial to the provision of valid data for these target
      compounds.


                                  E-30/SV"1                            OLM01.0

-------
      4.1      The  identification of all target compounds analyzed by the
               procedures described in Ex. D PEST is based on the use of
               absolute  retention time.  The mean retention time of each
               target  compound, or each peak in a multicomponent target
               compound, is determined from the initial calibration standards,
               according to the procedures outlined in Ex. 0 PEST. Section
               III, paragraph  8.

      4.2      The  retention time window of each target compound peak is
               determined as described in Ex. D, Section 111, paragraph 8.4.

      4.3      The  retention time shifts of the surrogates are used to
               evaluate  the stability of the gas chromatographic system during
               analysis  of samples and standards.  The retention time of the
               surrogates must be within the retention time windows determined
               during  the initial calibration (Ex. D PEST, Section III,
               paragraphs 7.10 and 8.4.

      4.4      The  documentation includes Form VI PEST, Form VII PEST, Form
               VIII PEST, chromatograms and data system printouts for all
               standards for the Pesticide/Aroclor initial and continuing
               calibrations, on each instrument and GC column used for
               analysis.

5.    Analytical Sequence

      The standards and samples analyzed according  to  the procedures  in Ex.  D
      PEST Section III must be analyzed in a sequence  described in paragraphs
      5 and 6.   This sequence includes  requirements  that  apply to the initial
      and continuing calibrations,  as well as  to  the analysis  of samples.
      The documentation includes  Form VIII PEST.

6.    Blank Analysis

      Two types of blanks  are required  for analyses using the  procedures  in
      Ex. D PEST.   They are method blanks  and  instrument blanks.   A  third
      type of blank,  a sulfur clean up  blank,  may be required.

      6.1     A method blank  is a volume of a clean reference matrix
               (deionized distilled water for water samples, or purified
              sodium sulfate  for soil/sediment samples) that is carried
               through the entire analytical procedure.  The volume or weight
              of the reference matrix must be approximately equal to the
              volume or weight of samples associated with the blank.   The
              purpose of a method blank is to determine the levels of
              contamination associated with the processing and analysis of
              samples.

              6.1.1      The frequency of method blank extraction is
                         described in  Ex.  D PEST,  Section III,  paragraph
                         15.1.1.
                                E-31/PEST                      OLM01.2  1/91

-------
        6.1.2      The method blank must be analyzed on each GC column
                   and instrument used for the analysis of associated
                   samples.

        6.1.3      For the purposes of this protocol, an acceptable
                   method blank must meet the criteria in Ex. D PEST,
                   Section,  III, paragraph 15.1.2.

6.2     The instrument blank is a volume of clean solvent spiked with
        the surrogates and analyzed on each GC column and instrument
        used for sample analysis.  The purpose of the instrument blank
        is to determine the levels of contamination associated with the
        instrumental analysis itself, particularly with regard to the
        carry-over of analytes from standards or highly contaminated
        samples into other analyses.

        6.2.1      The frequency of instrument blank analysis is part
                   of the initial and continuing calibration
                   requirements described in Ex. D,  Section III,
                   paragraphs 5,6, and 7.

        6.2.2      For the purposes of this protocol, an acceptable
                   instrument blank must meet the criteria in Ex. D
                   PEST, Section, III, paragraph 15.3.3.

6.3     The sulfur clean up blank is a volume of clean solvent spiked
        with the surrogates and carried through the sulfur clean up and
        analysis steps.  The purpose of the sulfur clean up blank is to
        determine the levels of contamination associated with the
        separate sulfur clean up steps.

        6.3.1      The sulfur clean up blank is only required when all
                   the samples associated with a particular method
                   blank are not subjected to sulfur clean up, as
                   described in Ex. D PEST, Section III, paragraph
                   15.2.2.

        6.3.2      The sulfur clean up blank must be analyzed on all GC
                   column and instruments used for analysis of samples
                   that received sulfur clean up.

        6.3.3      For the purposes of this protocol, an acceptable
                   sulfur clean up blank must meet the criteria in Ex.
                   D PEST, Section, III, paragraph 15.2.3.
                           E-32/SV                             OLM01.0

-------
      6.4      If a method blank exceeds  the  limits for contamination above,
               the Contractor must  consider the  analytical system out of
               control.   The  source of  the contamination must be investigated
               and appropriate corrective actions taken and documented before
               further  sample analysis  proceeds.  The  requirements  for
               reextraction and reanalysis of associated samples are given  in
               Ex.  D PEST,  Section  III, paragraph 15.1.3.

      6.5      If an instrument blank exceeds the limits for contamination
               above, the Contractor must consider the analytical system out
               of control.  The source  of the contamination must be
               investigated and appropriate corrective actions taken and
               documented before further  sample  analysis proceeds.  The
               requirements for reanalysis of associated samples are given  in
               Ex.  D PEST,  Section  III, paragraph 15.3.4.

      6.6      If a sulfur clean up blank exceeds the  limits for contamination
               above, the Contractor must consider the analytical system out
               of control.  The source  of the contamination must be
               investigated and appropriate corrective actions taken and
               documented before further  sample  analysis proceeds.  The
               requirements for reextraction  and reanalysis of associated
               samples are  given in Ex. D PEST,  Section III, paragraph 15.2.4.

      6.7      The  documentation includes Form I PEST  for the analysis of each
               type  of blank;  Form  IV PEST, associating the samples and the
               method and sulfur clean  up blank; Form  VIII PEST, associating
               the  samples  and the  instrument blanks;  and chromatograms and
               GC/EC data system printouts for the analysis of each blank.

7.     Surrogate Recoveries

      The recoveries of the two surrogates  are  calculated from the  analysis
      on each GC column of each sample, blank,  matrix  spike  and matrix spike
      duplicate.   The purpose of the surrogates  is  to  evaluate the
      preparation and analysis of samples.

      7.1      The  surrogates  are added to each sample, blank, matrix spike,
               and matrix spike  duplicate prior  to extraction, at the
               concentrations  described in Ex. D PEST,  Section II,  paragraph
               4.9.4.

      7.2      The  recoveries  of the surrogates are calculated according to
               the procedures  in Ex. D  PEST,  Section III,  paragraph 13.6.

      7.3      The quality  control  limits for surrogate recovery, given in Ex.
               D  PEST,. Section  III, paragraph 13.6, are 60-150 percent.
               These limits are  only advisory, and no  further action by the
               laboratory is required if  the  limits are exceeded, however,
               frequent failures  to meet  the  limits for surrogate recovery
               warrant investigation by the laboratory, and any result in
               questions  from  the Agency.
                                E-33/PEST                      OLM01.2   1/91

-------
      7.4      The  documentation  includes Form II PEST, a chromatogram and a
               GC/EC  data  system  printout for the analysis of each sample,
               blank, matrix  spike, and matrix spike duplicate.

8.    Matrix Spike and Matrix Spike Duplicate Analysis

      In order to evaluate the effects of the sample matrix on the methods
      used for pesticide/Aroclor analyses,  the Agency has prescribed a
      mixture of pesticide/Aroclor target compounds to be spiked into two
      aliquots of a sample, and analyzed in accordance with the appropriate
      method.

      8.1      The  frequency  of matrix spike and matrix spike duplicate
               (MS/MSD) analysis  is described in Ex. D PEST, Section  III,
               paragraph 16.1.

      8.2      The  recoveries of  the matrix spike compounds are calculated
               according to the procedures in Ex. D PEST, Section III,
               paragraph 16.3.  The relative percent difference for each
               spiked analyte between the results of the matrix spike and the
               matrix spike duplicate are calculated according to the
               procedures  in  Ex.  D PEST, Section III, paragraph 16.3.

      8.3      The  quality control limits for recovery and relative percent
               difference  are given in Ex. D PEST, Section III, paragraph
               16.4.  These limits are only advisory at this time, and no
               further action is  required when the limits are exceeded.

      8.4      The  documentation  includes Form I PEST for both the MS and MSD
               analyses, Form III PEST, and chromatograms and a GC/EC data
               system printout for each analysis.

9.    Dilution of Samples, Matrix Spikes,  and Matrix Spike Duplicates

      If the on-column concentration of any sample exceeds the initial
      calibration range,  that sample  must be diluted and reanalyzed,  as
      described in Ex.  D PEST,  Section III,  paragraph 13.4.   Guidance in
      performing dilutions and exceptions are given in that paragraph,  and
      reiterated here.

      9.1      If the response is still above the high calibration point after
               the  dilution of 1:100,000, the Contractor shall contact the SMO
               immediately.

      9.2     Use  the results of the original analysis to determine  the
               approximate dilution factor required to get the largest analyte
              peak within the initial calibration range.

      9.3     The dilution factor chosen should keep the response of the
               largest analyte peak for a target compound in the upper half of
               the  initial calibration range of the instrument.
                                E-34/PEST                       OLM01.2   1/91

-------
9.4     Do not submit data for more than two analyses, i.e., the
        original sample and one dilution, or, if the pesticide/Aroclor
        screening procedure was employed, from the most concentrated
        dilution analyzed and one further dilution.

9.5     Do not dilute MS/MSD samples to get either spiked or non-spiked
        analytes within the calibration range.  If the sample from
        which the MS/MSD aliquots were taken contains high levels of
        the spiked analytes, calculate the concentration and recovery
        of the analytes from the undiluted analysis and note the
        problem in the SDG Narrative.
                          E-35/PEST                     OLM01.1  12/90

-------
                                  SECTION V
                      ANALYTICAL STANDARDS REQUIREMENTS
Overview
The U.S. Environmental Protection Agency will not supply analytical reference
standards either for direct analytical measurements or for the purpose of
traceability.  All contract laboratories will be required to prepare from
neat materials or purchase from private chemical supply houses those
standards necessary to successfully and accurately perform the analyses
required in this protocol.

A.   Preparation of Chemical Standards from the  Neat High Purity Bulk
     Material

     A laboratory may prepare their chemical standards from neat materials.
     Commercial sources for neat chemical standards pertaining to compounds
     listed on the Compound Target List are given in the Appendix C of the
     "Quality Assurance Materials Bank: Analytical Reference Standards"
     Seventh Edition,  January 1988.   Laboratories should obtain the highest
     purity possible when purchasing neat chemical standards;  standards
     purchased at less than 97% purity must be documented as to why a higher
     purity could not be obtained.

     1.   Neat chemical standards must be kept refrigerated when not being
          used in the  preparation of standard solutions.   Proper storage  of
          neat chemicals is essential in order to safeguard them from
          decomposition.

     2.   The purity of a compound  can sometimes  be misrepresented by a
          chemical supply house.   Since knowledge of purity is needed to
          calculate the concentration of solute  in a solution standard, it is
          the contract laboratory's responsibility to have analytical
          documentation ascertaining that the purity of each compound is
          correctly stated.   Purity confirmation,  when performed,  should  use
          either differential scanning calorimetry,  gas chromatography with
          flame ionization detection,  high performance liquid chromatography,
          infrared spectrometry,  or other appropriate techniques.   Use of two
          or more independant methods is recommended.   The correction factor
          for impurity when weighing neat materials in the preparation of
          solution standards is:

          Equation 1

                                         weight  of pure compound
          weight of impure compound  -    (percent purity/100)

          where "weight of pure compound" is that required to prepare a
          specific volume of a solution standard of a specified concentration.
                                     E-36                              OLM01.0

-------
     3.   Mis-identification of compounds  occasionally occurs and  it  is
          possible that a mislabeled compound may be received from a  chemical
          supply house.  It is the  contract  laboratory's responsibility  to
          have analytical documentation ascertaining that all compounds  used
          in the preparation of solution standards be correctly  identified.
          Identification confirmation,  when  performed, should use  , gas
          chromatographic/mass spectrometry  analysis on at  least two  different
          analytical columns,  or other  appropriate techniques.

     4.   Calculate the weight of material to be weighed out for a specified
          volume taking into account the purity of the compound  and the
          desired concentration.  A second person must verify the  accuracy  of
          the calculations.   Check  balances  for accuracy with a  set of
          standard weights.   All weighing  should be performed on an analytical
          balance to the nearest 0.1 mg and  verified by a second person.  The
          solvent used to dissolve  the  solute should be compatible with  the
          protocol in which the standard is  to  be used; the solute should be
          soluble, stable,  and nonreactive with the solvent.  In the  case of a
          multicomponent solution,  the  components must not  react with each
          other.

     5.   Transfer the solute to a  volumetric flask and dilute to  the
          specified solution volume with solvent after ensuring  dissolution of
          the solute in the  solvent.  Sonication or warming may  be performed
          to promote dissolution of the solute.  This solution is  to  be  called
          the primary standard and  all  subsequent dilutions must be traceable
          back to the primary standard.

     6.   Log notebooks are  to be kept  for all  weighing and dilutions.   All
          subsequent dilutions from the primary standard and the calculations
          for determining their concentrations  are to be recorded  and verified
          by a second person.   All  solution  standards are to be  refrigerated
          when not in use.   All solution standards are to be clearly  labeled
          as to the identity of the compound or compounds,  concentration, date
          prepared,  solvent,  and initials  of the preparer.

B.   Purchase of chemical standards already  in  solution

     Solutions of analytical reference  standards can be purchased  by
     Contractors provided they meet the following criteria:

     1.   Laboratories must  maintain the following documentation to verify  the
          integrity of the  standard solutions they purchase:

          a.    mass  spectral  identification  confirmation of the  neat  material

          b.    purity confirmation  of the  neat  material

          c.    chromatographic and  quantitative  documentation that the
               solution standard  was  QC  checked according to the following
               section
                                     E-37                              OLM01.0

-------
The Contractor must purchase standards for which the quality is
demonstrated statistically and analytically by a method of the
supplier's choice.  One way this can be demonstrated is to prepare
and analyze three solutions; a high standard, a low standard, and a
standard at the target concentration (see parts a and b below) .  The
supplier must then demonstrate that the analytical results for the
high standard and low standard are consistent with the difference in
theoretical concentrations.  This is done by the Student's t-test in
part "d".  If this is achieved, the supplier must then demonstrate
that the concentration of the target standard lies midway between
the concentrations of the low and high standards .  This is done by
the Student's t-test in part e.  Thus the standard is certified to
be within 10 percent of the target concentration.

If the procedure above is used, the supplier must document that the
following have been achieved:

a.   Two solutions of identical concentration must be prepared
     independently from neat materials.   An aliquot of the first
     solution must be diluted to the intended concentration (the
     "target standard").   One aliquot is taken from the second
     solution and diluted to a concentration ten percent greater
     than the target standard.  This is called the "high standard".
     One further aliquot is taken from the second solution and
     diluted to a concentration 10 percent less that the target
     standard.  This is called the "low standard".

b.   Six replicate analyses of each standard (a total of 18
     analyses) must be performed in the following sequence: low
     standard, target,  high standard,  low standard,  target standard,
     high standard,  . . .

c.   The mean and variance of the six results for each solution must
     be calculated.

Equation 2
     MEAN  -  (YX + Y2  + ¥3 + Y4 + Y5  + Y6 )/6

Equation 3           222222           2
     VARIANCE  -  (Y]/  + Y2Z + Y3  + Y4  + Y5  + Y6  ' (6*MEAN)/)/5
     The values Y]_, Y2, Y3 ..... represent the results of the six
     analyses of each standard.  The means of the low, target, and
     high standards are designated MI, M2, and M3 ,  respectively.
     The variances of the low, target, and high standards are
     designated V]_, V2, and V3, respectively.  Additionally, a
     pooled variance, Vp, is calculated.

Equation 4
     Vp -  (V!/(0.81) + V2 + V3 /(1.21»/3

     If the square root of Vp is less than one percent of M2, then
     M2  /10.000 is to be used as the value of Vp in all subsequent
     calculations.

                            E-38                             OLM01.0

-------
     d.   The test statistic must be calculated:

          Equation 5
          TEST STATISTIC  -  |^3 /l.l) - (M! /0.9)|/(Vp /3)°'5

          If the test statistic exceeds 2.13 then the supplier has failed
          to demonstrate a twenty percent difference between the high and
          low standards.  In such a case, the standards are not
          acceptable.

     e.   The test statistic must be calculated:

          Equation 6
          TEST STATISTIC  -  |M2 - (Mi /I.8) - (M3 /2.2)|/(Vp /4)°'5

          If the test statistic exceeds 2.13, the supplier has failed to
          demonstrate that the target standard concentration is midway
          between the high and low standards.  In such a case, the
          standards are not acceptable.

     f.   The 95 percent confidence intervals for the mean result of each
          standard must be calculated:

          Equation 7
          Interval for Low Standard  -  M! ± (2.13)(Vp /6)°'5

          Equation 8
          Interval for Target Standard  -  M2 ± (2.13)(Vp /6)°'5

          Equation 9
          Interval for High Standard  -  M3 ± (2.13)(Vp /6)°'5

          These intervals must not overlap.  If overlap is observed, then
          the supplier has failed to demonstrate the ability to
          discriminate the 10 percent difference in concentrations.   In
          such a case, the standards are not acceptable.

 In any event,  the  laboratory  is  responsible for  the quality of the
 standards employed for analyses  under this contract.

Requesting Standards  From the EPA Standards  Repository

Solutions of analytical  reference materials  can be ordered from  the U.S.
EPA Chemical Standards Repository, depending on availability.  The
Contractor can place  an  order for standards  only after demonstrating
that these standards  are not available from  commercial vendors either in
solution or as a neat material.

Documentation of the  Verification and  Preparation of Chemical Standards

It is the responsibility of each laboratory  to maintain the necessary
documentation to show that the chemical  standards they have used in the
performance of CLP analysis conform to the  requirements previously
listed.  Weighing logbooks, calculations, chromatograms, mass spectra,

                                E-39                              OLM01.0

-------
etc, whether produced by the laboratory or purchased from chemical
supply houses, must be maintained by the laboratory and may be subject
to review during On-Site inspection visits.  In those cases where the
documentation is supportive of the analytical results of data packages
sent to EPA, such documentation is to be kept on file by the
laboratories for a period of one year.

Upon request by the Technical Project Officer or Administrative Project
Officer, the Contractor shall submit their most recent previous year's
documentation (12 months) for the verification and preparation of
chemical standards within 14 days of the receipt of request to the
recipients he/she designates.

The Agency may generate a report discussing deficiencies in the
Contractor's documentation for the verification and preparation of
chemical standards or may discuss the deficiencies during an On-Site
laboratory evaluation.  In a detailed letter to the Technical Project
Officer, Administrative Project Officer, and EMSL-LV, the Contractor
shall address the deficiencies and the subsequent corrective action
implemented by the Contractor to correct the deficiencies within 14 days
of receipt of the report or the On-Site laboratory evaluation.  An
alternate delivery schedule may be proposed by the Contractor, but it is
the sole decision of the Agency, represented either by the Technical
Project Officer or Administrative Project Officer, to approve or
disapprove the alternate delivery schedule.  If an alternate delivery
schedule is proposed, the Contractor shall describe in a letter to the
Technical Project Officer,  Administrative Project Officer, and the
Contracting Officer why he/she is unable to meet the delivery schedule
listed in this section.  The Technical Project Officer/Administrative
Project Officer will not grant an extension for greater than 14 days for
the Contractor's response letter to the standards documentation report.
The Contractor shall proceed and not assume that an extension will be
granted until so notified by the TPO and/or APO.

If new SOPs are required to be written or SOPs are required to be
amended because of the deficiencies and the subsequent corrective action
implemented by the Contractor, the Contractor shall write/amend and
submit the SOPs per the requirements listed in Exhibit E, Section III.

If the Contractor fails to adhere to the requirements listed in Section
V, a Contractor may expect, but the Agency is not limited to the
following actions:  reduction of number of samples sent under the
contract,  suspension of sample shipment to Contractor, GC/MS tape audit,
data package audit,  an On-Site laboratory evaluation, a remedial
laboratory evaluation sample, and/or contract sanctions, such as a Cure
Notice.
                              E-40                         OLM01.2  1/91

-------
                                  SECTION VI
                         CONTRACT  COMPLIANCE SCREENING
Contract Compliance Screening (CCS) is one aspect of the Government's
contractual right of inspection of analytical data. CCS examines the
Contractor's adherence to the contract requirements based on the sample data
package delivered to the Agency.

CCS is performed by the Sample Management Office (SMO) under the direction of
the EPA.  To assure a uniform review, a set of standardized procedures have
been developed to evaluate the sample data package submitted by a Contractor
against the technical and completeness requirements of the contract.

CCS results are mailed to the Contractor and all other data recipients.  The
Contractor has a period of time to correct deficiencies.  The Contractor must
send all corrections to the Regional Client, EMSL/LV, and SMO.

CCS results are used in conjunction with other information to measure overall
Contractor performance and to take appropriate actions to correct
deficiencies in performance.

The Agency may generate a CCS trend report which summarizes CCS results over
a given period of time.  The Agency may send the CCS trend report or discuss
the CCS trend report during an On-Site laboratory evaluation.  In a detailed
letter to the Technical Project Officer and Administrative Project Officer,
the Contractor shall address the deficiencies and the subsequent corrective
action implemented by the Contractor to correct the deficiencies within 14
days of receipt of the report or the On-Site laboratory evaluation.  An
alterante delivery schedule may be proposed by the Contractor, but it is the
sole decision of the Agency, represented by the Technical Project Officer or
Administrative Project Officer to approve or disprove the alternate delivery
schedule.  If an alternate delivery schedule is proposed, the Contractor
shall describe in a letter to the Technical project Officer, Administrative
Project Officer, and Contracting Officer why he/she is unable to meet the
delivery schedule listed in this section.  The Technical Project Officer will
not grant an extension for greater than 14 days for the Contractor's response
to the CCS trend report.

If new SOPs are required to be written or SOPs are required to be amended
because of the deficiencies and the subsequent corrective action implemented
by the Contractor, the Contractor shall write/amend and submit the SOPs per
the requirements listed in Exhibit E, Section III.

If the Contractor fails to adhere to the requirements listed in Section VI,
the Contractor may expect, but the Agency is not limited to the following
actions: reduction of number of samples sent under the contract, suspension
of sample shipment to the Contractor, GC/MS tape audit, data package audit,
an On-Site laboratory evaluation, a remedial laboratory evaluation sample,
and/or contract sanctions, such as a Cure Notice.
                                   E-41                         OLM01.2  1/91

-------
                                 SECTION VII
                             REGIONAL DATA REVIEW
Contract laboratory data are generated to meet the specific needs of the
Regions. In order to verify the useability of data for the intended purpose,
each Region reviews data from the perspective of end-user, based upon
functional aspects of data quality.  General guidelines for data review have
been developed jointly by the Region and the National Program Office.  Each
Region uses these guidelines as the basis for data evaluation.  Individual
Regions may augment the basic guideline review process with additional review
based on Region-specific or site-specific concerns.  Regional reviews, like
the sites under investigation, vary based on the nature .of the problems under
investigation and the Regional response appropriate to the specific
circumstances.

Regional data reviews, relating useability of the data to a specific site,
are part of the collective assessment process.  They complement the review
done at the Sample Management Office, which is designed to identify
contractual discrepancies and the review done at EMSL/LV which is designed to
evaluate Contractor and method performance.  These individual evaluations are
integrated into a collective review that is necessary for program and
laboratory administration and management and may be used to take appropriate
action to correct deficiencies in the Contractor's performance.
                                   E-42                               OLM01.0

-------
                                 SECTION VIII
                        LABORATORY EVALUATION SAMPLES
Although intralaboratory QC may demonstrate Contractor and method performance
that can be tracked over time, an external performance evaluation program is
an essential feature of a QA program.  As a means of measuring Contractor and
method performance, Contractors participate in interlaboratory comparison
studies conducted by the EPA.  Results from the analysis of these laboratory
evaluation samples, also referred to as performance evaluation (PE) samples,
will be used by the EPA to verify the Contractor's continuing ability to
produce acceptable analytical data. The results are also used to assess the
precision and accuracy of the analytical methods for specific analytes.

Sample sets may be provided to participating Contractors as frequently as on
an SDG-by-SDG basis as a recognizable QC sample of known composition; as a
recognizable QC sample of unknown composition; or not recognizable as a QC
material.  The laboratory evaluation samples may be sent either by the
Regional client or the National Program Office, and may be used for contract
action.

Contractors are required to analyze the samples and return the data package
and all raw data within the contract required turnaround time.

At a minimum, the results are evaluated for compound identification,
quantification, and sample contamination.  Confidence intervals for the
quantification of target compounds are based on reported values using
population statistics.  EPA may adjust the scores on any given laboraotry
evaluation sample to compensate for unanticipated difficulties with a
particular sample.  Normally, a fraction of the compounds spiked into the
sample are no't specifically listed in the contract.  Contractors are required
to use the NIST/EPA/MSDC mass spectral library to .tentatively identify a
maximum number of non-target compounds in each fraction that are present
above a minimal response.  Tentative identification of these compounds, based
on contractually described spectral interpretation procedures, is evaluated
and integrated into the evaluation process.

A Contractor's results on the laboratory evaluation samples will determine
the Contractor's performance as follows:

1.   Acceptable,  No Response Required (Score greater than or equal to 90
     percent):

     Data meets most or all of the scoring criteria.   No response is
     required.

2.   Acceptable,  Response Explaining Deficiency(ies)  Required (Score greater
     than or equal to 75 percent but less than 90 percent):

     Deficiencies exist in the Contractor's performance.
                                   E-43                               OLM01.0

-------
     Within 14 days of receipt of notification from EPA,  the  Contractor  shall
     describe the deficiency(ies) and the action(s)  taken to  correct  the
     deficiency(ies) in a letter to the Administrative  Project  Officer,  the
     Technical Project Officer and EMSL/LV.

     An alternate delivery schedule may be proposed by  the Contractor, but it
     is the sole decision of the Agency,  represented either by  the Technical
     Project Officer or Administrative Project Officer,  to approve of
     disapprove the alternate delivery schedule.   If an alternate delivery
     schedule  is proposed,  the Contractor shall  describe in  a  letter to the
     Technical Project Officer,  Administrative Project  Officer,  and the
     Contracting Officer why he/she is unable  to  meet the delivery schedule
     listed in this section.   The Technical  Project Officer /Administrative
     Project Officer will not grant an extension  for greater  than 14  days for
     the Contractor's response letter to the laboratory evaluation sample
     report.   The Contractor shall proceed and not assume that  an extension
     will be granted until so notified by the  TPO and/or APO.

     If new SOPs are required to be written  or SOPs  are  required to be
     amended because of the deficiencies and the  subsequent corrective action
     implemented by the Contractor, the Contractor shall write/amend  and
     submit the SOPs per the requirements listed  in Exhibit E,  Section III.

3.    Unacceptable Performance,  Response Explaining Deficiency(ies)  Required
     (Score less than 75 percent):

     Deficiencies exist in the Contractor's  performance  to the  extent that
     the National Program Office has determined that the Contractor has  not
     demonstrated the capability to meet the contract requirements.

     Within 14 days of receipt of notification from EPA,  the  Contractor  shall
     describe the deficiency(ies) and the action(s)  taken to  correct  the
     deficiency(ies) in a letter to the Administrative  Project  Officer,  the
     Technical Project Officer and EMSL/LV.

     An alternate delivery schedule may be proposed by  the Contractor, but it
     is the sole decision of the Agency,  represented either by  the Technical
     Project Officer or Administrative Project Officer,  to approve of
     disapprove the alternate delivery schedule.   If an  alternate delivery
     schedule  is proposed,  the Contractor shall  describe in  a  letter to the
     Technical Project Officer,  Administrative Project Officer,  and the
     Contracting Officer why he/she is unable  to  meet the delivery schedule
     listed in this section.   The Technical  Project  Officer /Administrative
     Project Officer will not grant an extension  for greater  than 14  days for
     the Contractor's response letter to the laboratory  evaluation sample
     report.

     If new SOPs are required to be written  or SOPs  are  required to be
     amended because of the  deficiencies  and the  subsequent corrective action
     implemented by the Contractor,  the Contractor shall write/amend and
     submit the SOPs per the  requirements listed  in Exhibit E,  Section III.
                                   E-44                        OLM01.2   1/91

-------
The Contractor shall be notified by the Technical Project Officer or
Administrative Project Officer concerning the remedy for their
unacceptable performance.  A Contractor may expect, but the Agency is
not limited to, the following actions: reduction of the number of
samples sent under the contract, suspension of sample shipment to the
Contractor, an On-Site laboratory evaluation, GC/MS tape audit, data
package audit, remedial laboratory evaluation sample, and/or a contract
sanction, such as a Cure Notice.

Note:  A Contractor's prompt response demonstrating that corrective
actions have been taken to ensure the Contractor's capability to meet
contract requirements may facilitate continuation of full sample
delivery.

If the Contractor fails to adhere to the requirements listed in Section
VIII, a Contractor may expect, but the Agency is not limited to the
following actions:  reduction in the number of samples sent under the
contract, suspension of sample shipment to the Contractor, an On-Site
laboratory evaluation, GC/MS tape audit, data package audit, a remedial
laboratory evaluation sample and/or contract sanctions, such as a Cure
Notice.
                              E-45                         OLM01.2  1/91

-------
                                  SECTION IX
                               GC/MS  TAPE AUDITS
Periodically, EPA requests from Contractors the GC/MS magnetic tapes
corresponding to a specific case in order to accomplish tape audits.
Generally, tape submissions and audits are requested for the following
reasons:

o   Program overview

o   Indication of data quality problems from EMSL/LV,  SMO,  or Regional data
    reviews

o   Support for On-Site audits
o   Specific Regional requests

Depending upon the reason for an audit, the tapes from a recent case, a
specific case, or a laboratory evaluation sample may be requested.  Tape
audits provide a mechanism to assess adherence to contractual requirements
and to ensure the consistency of data reported on the hardcopy/floppy
diskettes with that generated on the GC/MS tapes.  This function provides
external monitoring of Program QC requirements and checks adherence of the
Contractor to internal QA procedures.  In addition, tape audits enable EPA to
evaluate the utility, precision, and accuracy of the analytical methods.

The Contractor must store all raw and processed GC/MS data on magnetic tape,
in appropriate instrument manufacturer's format.  This tape must include data
for samples, blanks, matrix spikes, matrix spike duplicates, initial
calibrations, continuing calibrations, BFB and DFTPP, as well as all
laboratory-generated spectral libraries and quantitation reports required to
generate the data package.  The Contractor shall maintain a written reference
logbook of tape files to EPA sample number, calibration data, standards,
blanks, matrix spikes, and matrix spike duplicates.  The logbook should
include EPA sample numbers and standard and blank ID's, identified by Case
and Sample Delivery Group.

The Contractor is required to retain the GC/MS tapes for 365 days after data
submission.

When submitting GC/MS tapes to the Agency, the following materials must be
delivered in response to the request:

     1.  All associated raw data files for samples,  blanks,  matrix spikes,
         matrix  spike duplicates,  initial and continuing calibration
         standards,  and instrument performance check solutions (BFB and
         DFTPP).

     2.  All processed data files and quantitation output files  associated
         with the  raw data files described above.

     3.  All associated identifications and calculation files used to
         generate  the data submitted in the data package.

                                   E-46                         OLM01.2  1/91

-------
     4.   All laboratory-generated mass spectral library files  (NIST/EPA/MSDC
          library not required).

     5.   A copy of the Contractor's written reference logbook  relating tape
          files to EPA Sample Number, calibration data,  standards,  blanks,
          matrix spikes,  and matrix spike duplicates.   The logbook  must
          include EPA Sample Numbers and Lab File identifiers for all samples,
          blanks, and standards,  identified by Case and SDG.

The laboratory must also provide a statement attesting to the completeness of
the GC/MS data tape submission, signed and dated by the Laboratory Manager.
This statement must be part of a cover sheet that includes the following
information relevant to the data tape submission:

     1.   Laboratory name

     2.   Date of submission

     3.   Case Number

     4.   SDG Number

     5.   GC/MS make and model number

     6.   Software version

     7.   Disk drive type (e.g.  CDC,  PRIAM,  etc.)

     8.   File transfer method (e.g.  DSD,  DTD,  FTP,  Aquarius, etc.)

     9.   Names and telephone numbers of two laboratory contacts  for further
          information regarding the submission.

Submission of the GC/MS tape:

Upon request of the Administrative Project Officer or EMSL/LV,  the Contractor
shall send the required GC/MS tapes and all necessary documentation to
EMSL/LV within seven days of notification.  An alternate delivery schedule
may be proposed by the Contractor, but it is the sole decision of the Agency,
represented either by the Technical Project Officer or Administrative Project
Officer, to approve or disapprove the alternate delivery schedule.   If an
alternate delivery schedule is proposed, the Contractor shall describe in a
letter to the Technical Project Officer, Administrative Project Officer, and
the Contracting Officer why he/she is unable to meet the delivery schedule
listed in this section.  The Technical Project Officer/Administrative Project
Officer will not grant an extension for greater than seven days for
submission of the GC/MS tape.  The Contractor shall proceed and not assume
that an extension will be granted until so notified by the TPO and/or APO.
                                   E-47                         OLM01.2  1/91

-------
Responding to the GC/MS tape audit report:

After completion of the GC/MS tape audit, the Agency may send a copy of the
GC/MS tape audit report to the Contractor or may discuss the GC/MS tape audit
report on an On-Site laboratory evaluation.  In a detailed letter to the
Technical Project Officer, Administrative Project Officer, and EMSL/LV, the
Contractor shall discuss the corrective actions implemented to resolve the
deficiencies listed in the GC/MS tape audit report within 14 days of receipt
of the report.  An alternate delivery schedule may be proposed by the
Contractor, but it is the sole decision of the Agency, represented either by
the Technical Project Officer or Administrative Project Officer, to approve
or disapprove the alternate delivery schedule.  If an alternate delivery
schedule is proposed, the Contractor shall describe in a letter to the
Technical Project Officer, Administrative Project Officer, and the
Contracting Officer why he/she is unable to meet the delivery schedule listed
in this section.  The Technical Project Officer/Administrative Project
Officer will not grant an extension for greater than 14 days for the
Contractor's response letter to the GC/MS tape report.  The Contractor shall
proceed and not assume that an extension will be granted until so notified by
the TPO and/or APO.

If new SOPs are required to be written or SOPs are required to be amended
because of the deficiencies and the subsequent corrective action implemented
by the Contractor, the Contractor shall write/amend and submit the SOPs per
the requirements listed in Exhibit E, Section III.

Corrective actions

If the Contractor fails to adhere to the requirements listed in Section IX,
the Contractor may expect, but the Agency is not limited to the following
actions: reduction in the number of samples sent under the contract,
suspension of sample shipment to the Contractor, an On-Site laboratory
evaluation, GC/MS tape audit, data package audit, remedial laboratory
evaluation sample, and/or contract sanctions, such as a Cure Notice.
                                   E-48                         OLM01.2  1/91

-------
                                  SECTION X

                              DATA PACKAGE AUDITS

Data package audits are performed by the Agency for program overview and
specific Regional concerns.   Standardized procedures have been established to
assure uniformity of the auditing process.  Data .packages are periodically
selected from recently received cases.  They are evaluated for the technical
quality of hardcopy raw data, quality assurance, and the adherence to
contractual requirements.  This function provides external monitoring of
program QC requirements.

Data package audits are used  to assess the technical quality of the data and
evaluate overall laboratory performance.  It provides the Agency with an in-
depth inspection and evaluation of the Case data package with regard to
achieving QA/QC acceptability.  A thorough review of the raw data is
completed including:  a check of instrument printouts, quantitations reports,
chromatograms, spectra, library searches and other documentation for
deviations from the contractual requirements,  a check for transcription and
calculation errors, a review  of the qualifications of the laboratory
personnel involved with the Case, and a review of all current SOPs on file.

Responding to the data package audit report:

After completion of the data  package audit, the Agency may send a copy of the
data package audit report to  the Contractor or may discuss the data package
audit report on an On-Site laboratory evaluation.  In a detailed letter to
the Technical Project Officer, Administrative Project Officer, and EMSL/LV,
the Contractor shall discuss  the corrective actions implemented to resolve
the deficiencies listed in the data package audit report within 14 days of
receipt of the report.  An alternate delivery schedule may be proposed by the
Contractor, but it is the sole decision of the Agency, represented either by
the Technical Project Officer or Administrative Project Officer, to approve
or disapprove the alternate delivery schedule.  If an alternate delivery
schedule is proposed, the Contractor shall describe in a letter to the
Technical Project Officer, Administrative Project Officer, and the
Contracting Officer, why he/she is unable to meet the delivery schedule
listed in this section.  The  Technical Project Officer/Administrative Project
Officer will not grant an extension for greater than 14 days for the
Contractor's response letter  to the data package report.  The Contractor
shall proceed and not assume  that an extension will be granted until so
notified by the TPO and/or APO.

If new SOPs are required to be written or SOPs are required to be amended
because of the deficiencies and the subsequent corrective action implemented
by the Contractor, the Contractor shall write/amend and submit the SOPs per
the requirements listed in Exhibit E, Section III.
                                   E-49                         OLM01.2  1/91

-------
Corrective Actions

If the Contractor fails to adhere to the requirements listed in Section X,
the Contractor may expect, but the Agency is not limited to the following
actions:  reduction in the numbers of samples sent under the contract,
suspension of sample shipment to the Contractor, an On-Site laboratory
evaluation, GC/MS tape audit, data package audit,, remedial laboratory
evaluation sample, and/or contract sanctions, such as a Cure Notice.
                                   E-50                         OLM01.2  1/91

-------
                                  SECTION XI
                        ON-SITE LABORATORY EVALUATIONS
At a frequency dictated by a contract laboratory's performance, the
Administrative Project Officer, Technical Project Officer or their authorized
representative will conduct an On-Site laboratory evaluation.  On-site
laboratory evaluations are carried out to monitor the Contractor's ability to
meet selected terms and conditions specified in the contract.  The evaluation
process incorporates two separate categories:  Quality Assurance Evaluation,
and an Evidentiary Audit.

A.   Quality Assurance On-Site Evaluation

     o  Quality  assurance  evaluators  inspect the Contractor's  facilities  to
        verify the  adequacy and maintenance  of  instrumentation, the
        continuity  of  personnel meeting  experience or education requirements,
        and  the  acceptable performance of analytical and QC  procedures.   The
        Contractor  should  expect  that items  to  be monitored  will include, but
        not  be limited to  the  following  items.
     o  Size and appearance of the  facility
     o  Quantity, age, availability,  scheduled  maintenance and performance of
        instrumentation
     o  Availability,  appropriateness, and utilization of the  QAP  and SOPs
     o  Staff qualifications,  experience, and personnel training programs
     o  Reagents, standards, and  sample  storage facilities
     o  Standard preparation logbooks and raw data
     o  Bench sheets and analytical logbook  maintenance and  review
     o  Review of the  Contractor's  sample analysis/data package
        inspection/data management  procedures
     Prior to an On-Site  evaluation, various  documentation pertaining to
     performance  of the specific  Contractor is  integrated in a profile
     package  for  discussion during the evaluation.   Items  that may be
     included are previous  On-Site reports,  laboratory evaluation sample
     scores,  Regional review of data,  Regional  QA materials,  GO/MS  tape audit
     reports, data audit  reports,  results of  CCS,  and date trend reports.

B.   Evidentiary Audit

     Evidence auditors  conduct an On-Site laboratory evaluation to determine
     if laboratory policies and procedures  are  in place  to satisfy evidence
     handling requirements  as  stated in  Exhibit F.   The  evidence audit is
     comprised of the following three  activities:
                                   E-51                         OLM01.2  1/91

-------
     1.   Procedural Audit

          The procedural audit consists of review and examination of actual
          standard operating procedures and accompanying documentation for the
          following laboratory operations:   sample receiving,  sample storage,
          sample identification,  sample security,  sample tracking (from
          receipt to completion of analysis)  and analytical project file
          organization and assembly.

     2.   Written SOPs Audit

          The written SOPs audit consists of review and examination of the
          written SOPs to determine if they are accurate and complete for the
          following laboratory operations:   sample receiving,  sample storage,
          sample identification,  sample security,  sample tracking (from
          receipt to completion of analysis)  and analytical project file
          organization and assembly.

     3.   Analytical Project File Evidence Audit

          The analytical project file evidence audit consists  of review and
          examination of the analytical project file documentation.   The
          auditors review the files to determine:

          o  The  accuracy of  the document  inventory

          o  The  completeness  of the  file

          o  The  adequacy and  accuracy of  the  document numbering  system

          o  Traceability of sample activity

          o  Identification of activity  recorded  on  the documents

          o  Error correction  methods

C.   Discussion of the On-Site Team's Findings

     The quality assurance and evidentiary auditors discuss their findings
     with the Administrative Project Officer/Technical Project Officer prior
     to debriefing the Contractor.   During the debriefing,  the auditors
     present their findings and recommendations for corrective actions
     necessary to the Contractor personnel.

D.   Corrective Action Reports For Follow-Through to Quality Assurance and
     Evidentiary Audit Reports

     On-site laboratory evaluation:

     Following an On-Site laboratory evaluation, quality assurance and/or
     evidentiary audit reports which discuss deficiencies found during the
     On-Site evaluation may be sent to the Contractor.  In a detailed letter,
     the Contractor shall discuss the corrective actions implemented to
     resolve the deficiencies discussed during the On-Site evaluation and
     discussed in the report(s) to the Technical Project Officer,
     Administrative Project Officer,  and EMSL/LV (response to quality
     assurance/technical report)  and NEIC (response to the evidentiary

                                   E-52                         OLM01.2  1/91

-------
report), within 14 days of receipt of the report.  An alternate delivery
schedule may be proposed by the Contractor, but it is the sole decision
of the Agency, represented either by the Technical Project Officer or
Administrative Project Officer, to approve or disapprove the alternate
delivery schedule.  If an alternate delivery schedule is proposed, the
Contractor shall describe in a letter to the Technical Project Officer,
Administrative Project Officer, and the Contracting Officer why he/she
is unable to meet the delivery schedule listed in this section.  The
Technical Project Officer/Administrative Project Officer will not grant
an extension for greater than 14 days for the Contractor's response
letter to the quality assurance and evidentiary audit report.  The
Contractor shall proceed and not assume that an extension will be
granted until so notified by the TPO and/or APO.

If new SOPs are required to be written or SOPs are required to be
amended because of the deficiencies and the subsequent corrective action
implemented by the Contractor, the Contractor shall write/amend and
submit the SOPs per the requirements listed in Exhibit E, Section III.

Corrective actions

If the Contractor fails to adhere to the requirements listed in Section
XI, the Contractor may expect, but the Agency is not limited to the
following actions:  reduction in the number of samples sent under the
contract, suspension of sample shipment to the Contractor, an On-Site
laboratory evaluation, GC/MS tape audit, data package audit, a remedial
laboratory evaluation sample,  and/or contract sanctions,  such as a Cure
Notice.
                              E-53                         OLM01.2  1/91

-------
                                 SECTION XII


                  QUALITY ASSURANCE AND DATA TREND ANALYSIS


Data submitted by laboratories are subject to review from several aspects:
compliance with contract-required QC,  useability, and full data package
evaluation.  Problems resulting from any of these reviews may determine the
need for a GC/MS tape audit, an On-Site laboratory evaluation and/or a
remedial laboratory evaluation sample.  In addition,  QC prescribed in the
methods provides information that is continually used by the Agency to assess
sample data quality, Contractor data quality and Program data quality via
data trend analysis.  Trend analysis is accomplished by entering data into a
computerized data base.  Statistical reports that evaluate specific anomalies
or disclose trends in many areas, including the following, are generated from
this data base:

o   Surrogate  Spike  Recovery

o   Laboratory Evaluation Sample
o   Blanks

o   GC/MS Instrument Performance  Checks  (BFB and DFTPP)
o   Initial and Continuing Calibration Data

o   Other QC and Method Parameters
Program-wide statistical results are used to rank laboratories in order to
observe the relative performance of each Contractor using a given protocol
against its peers.  The reports are also used to identify trends within
laboratories.   The results of many of these trends analyses are included in
overall evaluation of a Contractor's performance, and are reviewed to
determine if corrective action or an On-Site laboratory evaluation is
indicated in order to meet the QA/QC requirements of the contract.

Contractor performance over time is monitored using these trend analysis
techniques to detect departures of Contractor output from required or desired
levels of quality control, and to provide an early warning of Contractor
QA/QC problems which may not be apparent from the results of an individual
case.

As a further benefit to the Program, the data base provides the information
needed to establish performance-based criteria in updated analytical
protocols,  where advisory criteria has been previously used.  The vast
empirical data set produced by contract laboratories is carefully analyzed,
with the results augmenting theoretical and research-based performance
criteria.   The result is a continuously monitored set of quality control and
performance criteria specifications of what is routinely achievable and
expected of environmental chemistry laboratories in mass production analysis
of environmental samples.  This, in turn, assists the Agency in meeting its
objectives of obtaining data of known and documented quality.
                                   E-54                               OLM01.0

-------
                                 SECTION XIII
                               DATA MANAGEMENT
Data management procedures are defined as procedures specifying the
acquisition or entry, update, correction, deletion, storage and security of
computer readable data and files.  These procedures should be in written form
and contain a clear definition for all databases and files used to generate
or resubmit deliverables.  Key areas of concern include:  system organization
(including personnel and security), documentation operations, traceability
and quality control.

Data manually entered from hard-copy must be quality controlled and the error
rates estimated.  Systems should prevent entry of incorrect or out-of-range
data and alert data entry personnel of errors.  In addition, data entry error
rates must be estimated and recorded on a monthly basis by reentering a
statistical sample of the data entered and calculating discrepancy rates by
data element.

The record of changes in the form of corrections and updates to data
originally generated, submitted, and/or resubmitted must be documented to
allow traceablilty of updates.  Documentation must include the following for
each change:

o   Justification or rationale for the change.

o   Initials of the person making the  change or changes.   Data changes must be
    implemented and reviewed by a person or group independant of the source
    generating the deliverable.

o   Change documentation must be retained according to  the schedule of the
    original deliverable.

o   Resubmitted diskettes or other deliverables must be reinspected as a part
    of the laboratories'  internal inspection process prior to resubmission.
    The entire deliverable,  not just the changes,  must  be  inspected.

o   The Laboratory Manager must approve changes to originally submitted
    deliverables.

o   Documentation of data changes may  be requested by laboratory auditors.

Lifecycle management procedures must be applied to computer software systems
developed by the laboratory to be used to generate and edit contract
deliverables.  Such systems must be thoroughly tested and documented prior to
utilization.

o   A software test and acceptance plan including test  requirements,  test
    results  and acceptance criteria must be developed,  followed,  and available
    in written form.

o   System changes must not be made directly to production systems generating
    deliverables.   Changes must be made first to a development system and
    tested prior to implementation.
                                   E-55                               OLM01.0

-------
o   Each version of the production system will be  given  an  identification
    number,  date of installation,  date  of last operation and  archived.

o   System and operations documentation must be  developed and maintained for
    each system.   Documentation must include a users manual and  an  operations
    and maintenance manual.

Individual(s) responsible for the following functions must be identified:

o   System operation and maintenance including documentation  and training.

o   Database integrity,  including  data  entry, data updating and  quality
    control.

o   Data and system security,  backup and archiving.
                                   E-56                               OLM01.0

-------
             EXHIBIT F
CHAIN-OF-CUSTODY, DOCUMENT CONTROL,
 AND STANDARD OPERATING PROCEDURES
              F-l                                OLM01.0

-------
1.    SAMPLE CHAIN-OF-CUSTODY

      A sample is physical evidence collected from a facility or from the
      environment.  Controlling evidence is an essential part of the
      hazardous waste investigation effort.  To accomplish this, Contractors
      are required to develop and implement the following sample
      identification, chain-of-custody, sample receiving, and sample tracking
      procedures.

1.1   Sample Identification

      To assure traceability of the samples while in possession of the
      Contractor, the Contractor shall have a specified method for
      maintaining identification of samples throughout the laboratory.

      Bach sample and sample preparation container shall be labeled with the
      EPA number or a unique laboratory identifier.  If a unique laboratory
      identifier is used, it shall be cross-referenced to the EPA number.

1.2   Chain-of-Custody Procedures

      Because of the nature of the data being collected, the custody of EPA
      samples must be traceable from the time the samples are collected until
      they are introduced as evidence in legal proceedings.  The Contractor
      shall have procedures ensuring that EPA sample custody is maintained
      and documented.  A sample is under custody if:

          o   It is in your possession, or

          o   It is in your view after being in your possession, or

          o   It was in your possession and you locked it up, or

          o   It is in a designated secure area.  (Secure areas shall be
             accessible only to authorized personnel.)

1.3   Sample Receiving Procedures

      1.3.1   The Contractor shall designate  a sample custodian responsible
              for receiving all samples.

      1.3.2   The Contractor shall designate  a representative to receive
              samples in the event that the sample  custodian is not
              available.

      1.3.3   The condition of the shipping containers  and sample bottles
              shall be inspected upon  receipt by the sample custodian or
              his/her representative.

      1.3.4   The condition of the custody seals (intact/not intact)  shall be
              inspected upon receipt by the sample  custodian or his/her
              repre sent at ive.

      1.3.5   The sample custodian or  his/her representative shall check for
              the presence or absence  of the  following  documents accompanying
              the sample shipment:

                                   F-2                               OLM01.0

-------
        o   Airbills or airbill stickers
        o   Custody seals
        o   EPA custody records
        o   EPA traffic reports or SAS packing lists
        o   Sample tags
1.3.6   The sample custodian or his/her representative shall sign and
        date all  forms  (e.g., custody records, traffic reports or
        packing lists,  and airbills) accompanying the samples at the
        time of sample  receipt.
1.3.7   The Contractor  shall contact the Sample Management Office (SMO)
        to resolve discrepancies and problems such as absent documents,
        conflicting information, broken custody seals, and
        unsatisfactory  sample condition (e.g., leaking sample bottle).
1.3.8   The Contractor  shall record the resolution of discrepancies and
        problems  on Telephone Contact Logs.
1.3.9   The following information shall be recorded on Form DC-1 (See
        Exhibit B) by the sample custodian or his/her representative as
        samples are received and inspected:
        o   Condition of the shipping container
        o   Presence or absence and condition of custody seals on
            shipping and/or sample containers
        o   Custody seal numbers,  when present
        o   Condition of the sample bottles
        o   Presence or absence of airbills or airbill stickers
        o   Airbill or airbill sticker numbers
        o   Presence or absence of EPA custody records
        o   Presence or absence of EPA traffic reports or SAS packing
            lists
        o   Presence or absence of sample  tags
        o   Sample tag identification numbers cross-referenced to the
            EPA sample numbers
        o   Verification of agreement or non-agreement of information
            recorded on shipping documents and sample containers
        o   Problems or discrepancies
                             F-3                               OLM01.0

-------
1.4   Sample Tracking Procedures

      The Contractor shall maintain records documenting all phases of sample
      handling from receipt to final analysis.

2.    DOCUMENT CONTROL PROCEDURES

      The goal of the laboratory document control program is to assure that
      all documents for a specified Sample Delivery Group (SDG) will be
      accounted for when the project is completed.  Accountable documents
      used by contract laboratories shall include, but not be limited to,
      logbooks, chain-of-custody records, sample work sheets, bench sheets,
      and other documents relating to the sample or sample analyses.  The
      following document control procedures have been established to assure
      that all laboratory records are assembled and stored for delivery to
      EPA or are available upon request from EPA prior to the delivery
      schedule.

2.1   Preprinted Laboratory Forms and Logbooks

      2.1.1   All documents produced by the Contractor which are directly
              related to the preparation and analysis of EPA samples shall
              become the property of the EPA and  shall be placed in the
              complete sample delivery group file (CSF).  All observations
              and results recorded by the laboratory but not on preprinted
              laboratory forms shall be entered into permanent laboratory
              logbooks.  When all data from a SDG is compiled, all original
              laboratory forms and copies of all  SDG-related logbook entries
              shall be included in the documentation package.

      2.1.2   The Contractor shall identify the activity recorded on all
              laboratory documents which are directly related to the
              preparation and analysis of EPA samples.

      2.1.3   Pre-printed laboratory forms shall  contain the name of the
              laboratory and be dated (month/day/year) and signed by the
              person responsible for performing the activity at the time an
              activity is performed.

      2.1.4   Logbook entries shall be dated (month/day/year) and signed by
              the person responsible for performing the activity at the time
              an activity is performed.

      2.1.5   Logbook entries shall be in chronological order.  Entries in
              logbooks, with the exception of instrument run logs and
              extraction logs, shall include only one SDG per page.

      2.1.6   Pages in both bound and unbound logbooks shall be sequentially
              numbered.
                                   F-4                               OLM01.0

-------
       2.1.7    Instrument run logs shall be maintained so as to enable a
               reconstruction of the run sequence of individual instruments.

               Because the laboratory must provide copies of the instrument
               run logs to EPA,  the laboratory may exercise the option of
               using only laboratory or EPA sample identification numbers in
               the logs for sample ID rather than.government agency or
               commercial client names to preserve the confidentiality of
               commercial clients.

       2.1.8    Corrections to supporting documents and raw data shall  be made
               by drawing a single line through the error and entering the
               correct information.  Corrections and additions to supporting
               documents and raw data shall be dated and initialed.  No
               information shall be obliterated or rendered unreadable.

               All notations shall be recorded in ink.

               Unused portions of documents shall be "z'd" out.

2.2    Consistency of Documentation

       The Contractor shall  assign a document control officer responsible  for
       the organization  and  assembly of the CSF.

       All copies of laboratory documents shall be complete and legible.

       Original documents which include information relating to more than  one
       SDG shall be filed in the CSF of the lowest SDG number.  The copy(s)
       shall be placed in the other CSF(s) and the Contractor shall record the
       following information on the copy(s) in red ink:

            "COPY

      ORIGINAL IS FILED IN CSF 	"
      The Contractor shall sign and date this addition to the copy(s).

      Before releasing analytical results, the document control officer shall
      assemble and cross-check the information on samples tags, custody
      records, lab bench sheets, personal and instrument logs, and other
      relevant deliverables to ensure that data pertaining to each particular
      sample or sample delivery group is consistent throughout the CSF.

2.3   Document Numbering and Inventory Procedure

      In order to provide document accountability of the completed analysis
      records, each item in the CSF shall be inventoried and assigned a
      serialized number as described in Exhibit B).
                                   F-5                                OLM01.0

-------
      All documents relevant to each sample delivery group,  including logbook
      pages, bench sheets, mass spectra,  chromatograms,  screening records,
      re-preparation records, re-analysis records,  records of failed or
      attempted analysis, custody records,  library research results, etc.
      shall be inventoried.

      The Document Control Officer (DCO)  shall be responsible for ensuring
      that all documents generated are placed in the CSF for inventory and
      are delivered to the appropriate EPA region or other receiver as
      designated by EPA.  The DCO shall place the sample tags in plastic bags
      in the file.

2.4   Storage of EPA Files

      The Contractor shall maintain EPA laboratory documents in a secure
      location.

2.5   Shipment of Deliverables

      The Contractor shall document shipment of deliverables packages to the
      recipients.  These shipments require custody seals on the containers
      placed such that they cannot be opened without damaging or breaking the
      seal.  The Contractor shall document what was sent,  to whom, the date,
      and the method (carrier) used.

3.    SPECIFICATIONS FOR WRITTEN STANDARD OPERATING PROCEDURES

      The Contractor shall have written standard operating procedures (SOPs)
      for receipt of samples, maintenance of custody, sample identification,
      sample storage, sample tracking, and assembly of completed data.
      An SOP is defined as a written narrative stepwise description of
      laboratory operating procedures including examples of laboratory
      documents.   The SOPs shall accurately describe the actual procedures
      used in the laboratory, and copies of the written SOPs shall be
      available to the appropriate laboratory personnel.  These procedures
      are necessary to ensure that analytical data produced under this
      contract are acceptable for use in EPA enforcement case preparation and
      litigation.  The Contractor's SOPs shall provide mechanisms and
      documentation to meet each of the following specifications and shall be
      used by EPA as the basis for laboratory evidence audits.

3.1   The Contractor shall have written SOPs describing the sample
      custodian's duties and responsibilities.

3.2   The Contractor shall have written SOPs for receiving and logging in of
      the samples.  The procedures shall include but not be limited to
      documenting the following information:

      3.2.1    Presence  or  absence of  EPA chain-of-custody forms

      3.2.2    Presence  or  absence of  airbills or airbill stickers
                                   F-6                          OLM01.5   4/91

-------
      3.2.3    Presence or absence of traffic reports or SAS packing lists

      3.2.4    Presence or absence of custody seals on shipping and/or sample
               containers and their condition

      3.2.5    Custody seal numbers, when present

      3.2.6    Airbill or airbill sticker numbers

      3.2.7    Presence or absence of sample tags

      3.2.8    Sample tag ID numbers

      3.2.9    Condition of the shipping container

      3.2.10   Condition of the sample bottles

      3.2.11   Verification of agreement or non-agreement of information on
               receiving documents and sample containers

      3.2.12   Resolution of problems or discrepancies with the SMO

      3.2.13   An explanation of any terms used by the laboratory to describe
               sample condition upon receipt (e.g., good, fine, OK)

3.3   The Contractor shall  have written SOPs for maintaining identification
      of EPA samples throughout the laboratory.

      If the Contractor assigns unique laboratory identifiers, written SOPs
      shall include a description of the method used to assign the unique
      laboratory identifier and shall include  a description of the document
      used to cross-reference the unique laboratory identifier to the EPA
      sample number.

      If the Contractor uses prefixes or suffixes in addition to sample
      identification numbers,  the written SOPs shall include their
      definitions.

3.4   The Contractor shall  have written SOPs describing all storage areas for
      samples in the laboratory.   The SOPs shall include a list of authorized
      personnel who have access or keys to secure storage areas.

3.5   The Contractor shall  have written SOPs describing the method by which
      the laboratory maintains samples under custody.

3.6   The Contractor shall  have written SOPs describing the method by which
      the laboratory maintains the security of any areas identified as
      secure.

3.7   The Contractor shall  have written SOPs for tracking the work performed
      on any particular samples.   The tracking SOP shall include:
                                   F-7                                OLM01.0

-------
          o  A description of the documents used to record sample receipt,
             sample storage, sample transfers, sample preparations, and
             sample analyses.

          o  A description of the documents used to record calibration and
             QA/QC laboratory work.

          o  Examples of document formats and laboratory documents used in
             the sample receipt, sample storage, sample transfer, and sample
             analyses.

          o  A narrative step-wise description of how documents are used to
             track samples.

3.8   The Contractor shall have written SOPs for organization and assembly of
      all documents relating to each SOG.   Documents shall be filed on a
      sample delivery group-specific basis.  The procedures shall ensure that
      all documents including logbook pages, sample tracking records,
      chromatographic charts, computer printouts, raw data summaries,
      correspondence,  and any other written documents having reference to the
      SDG are compiled in one location for submission to EPA.  The written
      SOPs shall include:

          o  A description of the numbering and inventory method.

          o  A description of the method used by the laboratory to verify
             consistency and completeness of the CSF.

          o  Procedures for the shipment of deliverables packages using
             custody seals.

4.    HANDLING OF CONFIDENTIAL INFORMATION

      A Contractor conducting work under this contract may receive EPA-
      designated confidential information from the agency.  Confidential
      information must be handled separately from other documentation
      developed under this contract.  To accomplish this, the following
      procedures for the handling of confidential information have been
      established.

4.1   All confidential documents shall be under the supervision of a
      designated document control officer (DCO).

4.2   Confidential Information

      Any samples or information received with a request of confidentiality
      shall be handled as 'confidential.*   A separate locked file shall be
      maintained to store this information and shall be segregated from other
      nonconfidential  information.  Data generated from confidential samples
      shall be treated as confidential.  Upon receipt of confidential
      information, the DCO will log these documents into a Confidential
      Inventory Log.   The information will then be available to authorized
      personnel but only after it has been signed out to that person by the
      DCO.  The documents shall be returned to the locked file at the
      conclusion of each working day.   Confidential information may not be
      reproduced except upon approval by the EPA Technical and Administrative
      Project Officer.  The DCO will enter all copies into the document
      control system described above.   In addition, this information may not

                                   F-8                                OLM01.0

-------
be disposed of except upon approval by the EPA project officer.  The
DCO shall remove and retain the cover page of any confidential
information disposed of for one year and shall keep a record on the
disposition in the Confidential Inventory Log.
                             F-9                               OLM01.0

-------
    EXHIBIT G
GLOSSARY OF TERMS
     G-l                               OLM01.0

-------
                               GLOSSARY OF  TERMS
ALIQUOT  - a measured portion of a sample taken for analysis.

ANALYSIS DATE/TIME  - the date and military time of the injection of the
sample, standard, or blank into the GC/MS or GC system.

BAR GRAPH SPECTRUM  - a plot of the mass-to-charge ratio (m/e) versus relative
intensity of the ion current.

BLANK  - see Method  Blank

4-BROMOFLUOROBENZENE (BFB) - compound chosen to establish mass spectral
instrument performance for volatile analyses.

CASE - a finite, usually predetermined number of samples collected over a
given  time period from a particular site. Case numbers are assigned by the
Sample Management Office.  A Case consists of one or more Sample Delivery
Groups.

CHARACTERIZATION -  a determination of the approximate concentration range of
compounds of interest used to choose the appropriate analytical protocol.

CONCENTRATION LEVEL (low or medium) - characterization of soil samples or
sample fractions as low concentration or medium concentration is made on the
basis of the laboratory's preliminary screen, not on the basis of information
entered on the Traffic Report by the sampler.

CONTINUING CALIBRATION - analytical standard run every 12 hours to verify the
calibration of the GC/MS system.

CONTINUOUS LIQUID-LIQUID EXTRACTION - used herein synonymously with the terms
continuous extraction,  continuous liquid extraction,  and liquid extraction.
This extraction technique involves boiling the extraction solvent in a flask
and condensing the  solvent above the aqueous sample.   The condensed solvent
drips through the sample, extracting the compounds of interest from the
aqueous phase.

DAY - unless otherwise specified, day shall mean calendar day.

DECAFLUOROTRIPHENYLPHOSPHINE (DFTPP) - compound chosen to establish mass
spectral instrument performance for semivolatile analysis.

EXTRACTABLE - a compound that can be partitioned into an organic solvent from
the sample matrix and is amenable to gas chromatography.  Extractables
include semivolatile (BNA) and pesticide/Aroclor compounds.

IN-HOUSE - at the Contractor's facility.

INITIAL CALIBRATION - analysis of analytical standards for a series of
different specified concentrations; used to define the linearity and dynamic
range of the response of the mass spectrometer or electron capture detector
to the target compounds.


                                    G-2                          OLM01.2  1/91

-------
INTERNAL STANDARDS - compounds added Co every standard, blank, matrix spike,
matrix spike duplicate, sample (for VOAs),  and sample extract (for
semivolatiles) at a known concentration, prior to analysis.  Internal
standards are used as the basis for quantitation of the target compounds.

LABORATORY - synonymous with Contractor as used herein.

m/z - Mass to charge ratio, synonymous with "m/e".

MATRIX - the predominant material of which the sample to be analyzed is
composed. For the purpose of this SOW, a sample matrix is either water or
soil/sediment. Matrix is not synonymous with phase (liquid or solid).

MATRIX SPIKE - aliquot of a matrix (water or soil) fortified (spiked) with
known quantities of specific compounds and subjected to the entire analytical
procedure in order to indicate the appropriateness of the method for the
matrix by measuring recovery.

MATRIX SPIKE DUPLICATE - a second aliquot of the same matrix as the matrix
spike (above) that is spiked in order to determine the precision of the
method.

METHOD BLANK (previously termed reagent blank) - an analytical control
consisting of all reagents, internal standards and surrogate standards, that
is carried through the entire analytical procedure.  The method blank is used
to define the level of laboratory background and reagent contamination.

NARRATIVE (SDG Narrative) - portion of the data package which includes
laboratory, contract, Case and sample number identification, and descriptive
documentation of any problems encountered in processing the samples, along
with corrective action taken and problem resolution.  Complete SDG Narrative
specifications are included in Exhibit B.

PERCENT DIFFERENCE (%D) - As used in this SOU and elsewhere to compare two
values, the percent difference indicates both the direction and the magnitude
of the comparison, i.e., the percent difference may be either negative,
positive, or zero.  (In contrast, see relative percent difference below).

PERCENT MOISTURE - an approximation of the amount of water in a soil/sediment
sample made by drying an aliquot of the sample at 105°C.  The percent
moisture determined in this manner also includes contributions from all
compounds that may volatilize at or below 105°C, including water.  Percent
moisture may be determined from decanted samples and from samples that are
not decanted.

PROTOCOL - describes the exact procedures to be followed with respect to
sample receipt and handling, analytical methods, data reporting and
deliverables, and document control.  Used synonymously with Statement of Work
(SOW).
                                   G-3                          OLM01.2  1/91

-------
 PURGE AND TRAP  (DEVICE) - analytical technique (device) used to isolate
 volatile  (purgeable) organics by stripping the compounds from water or soil
 by a stream of  inert gas, trapping the compounds on an adsorbent such as a
 porous polymer  trap, and thermally desorbing the trapped compounds onto the
 gas chromatographic column.

 REAGENT WATER - water in which an interferent is not observed at or above the
 minimum quantitation limit of the parameters of interest.

 RECONSTRUCTED ION CHROMATOGRAM (RIG) - a mass spectral graphical
 representation  of the separation achieved by a gas chromatograph;  a plot of
 total ion current versus retention time.

 RELATIVE PERCENT DIFFERENCE  (RPD) - As used in this SOW and elsewhere to
 compare two values, the relative percent difference is based on the mean of
 the two values, and is reported as an absolute value, i.e., always expressed
 as a positive number or zero.  (In contrast, see percent difference above).

 RELATIVE RESPONSE FACTOR (RRF) - a measure of the relative mass spectral
 response of an  analyte compared to its internal standard.  Relative Response
 Factors are determined by analysis of standards and are used in the
 calculation of  concentrations of analytes in samples.  RRF is determined by
 the following equation:

                A       f •
                AX      °LS
      RRF  =    —  x  —-
                Ais     cx

      Where

      A  = area of the characteristic ion measured

      C  = concentration

      is = internal standard

      x  = analyte of interest

RESOLUTION - also termed separation or percent resolution, the separation
between peaks on a chromatogram,  calculated by dividing the depth of the
valley between  the peaks by the peak height of the smaller peak being
resolved,  multipled by 100.
                                             Resolution = A x  1OO
                                                           B
                                    G-4                         OLM01.2  1/91

-------
SAMPLE - a portion of material to be analyzed that is contained in single or
multiple containers and identified by a unique sample number.

SAMPLE DELIVERY GROUP (SDG) - a unit within a single Case that is used to
identify a group of samples for delivery.  An SDG is a group of 20 or fewer
field samples within a Case, received over a period of up to 14 calendar days
(7 calendar days for 14-day data turnaround contracts).  Data from all
samples in an SDG are due concurrently.  A Sample Delivery Group is defined
by one of the following, whichever occurs first:

o   Case;  or

o   Each 20 field samples  within a Case;  or

o   Each 14-day calendar period (7-day calendar  period for 14-day data
    turnaround contracts)  during which field samples  in a Case  are received,
    beginning with receipt of the first sample in the Case or SDG.

Samples may be assigned to Sample Delivery Groups by matrix (i.e., all soils
in one SDG, all waters in another), at the discretion of the laboratory.

SAMPLE NUMBER (EPA Sample Number) - a unique identification number designated
by EPA for each sample.  The EPA sample number appears on the sample Traffic
Report which documents information on that sample.

SEMIVOLATILE COMPOUNDS - compounds amenable to analysis by extraction of the
sample with an organic solvent.  Used synonymously with Base/Neutral/Acid
(BNA) compounds.

SOIL - used herein synonymously with soil/sediment and sediment.

STANDARD ANALYSIS - an analytical determination made with known quantities of
target compounds; used to determine response factors.

SURROGATES (Surrogate Standard) - for semivolatiles and pesticides/Aroclors,
compounds added to every blank, sample, matrix spike, matrix spike duplicate,
and standard; used to evaluate analytical efficiency by measuring recovery.
Surrogates are brominated, fluorinated, or isotopically labelled compounds
not expected to be detected in environmental media.

SYSTEM MONITORING COMPOUNDS - compounds added to every blank, sample, matrix
spike, matrix spike duplicate, and standard for volatile analysis, and used
to evaluate the performance of the entire purge and trap-gas chromatograph-
mass spectrometer system.   These compounds are brominated or deuterated
compounds not expected to be detected in environmental media.

TARGET COMPOUND LIST (TCL) - a list of compounds designated by the Statement
of Work (Exhibit C) for analysis.

TENTATIVELY IDENTIFIED COMPOUNDS (TIC) - compounds detected in samples that
are not target compounds,  internal standards, system monitoring compounds, or
surrogates.  Up to 30 peaks (those greater than 10% of peak areas or heights
of nearest internal standards) are subjected to mass spectral library
searches for tentative identification.
                                    G-5                          OLM01.2  1/91

-------
TIME  - when required to record time on any deliverable item, time shall be
expressed as Military Time, i.e., a 24-hour clock.

TRAFFIC REPORT (TR) - an EPA sample identification form filled out by the
sampler, which accompanies the sample during shipment to the laboratory and
which documents sample condition and receipt by the laboratory.

TWELVE-HOUR TIME PERIOD - The twelve (12) hour time period for GC/MS system
instrument performance check, standards calibration (initial or continuing
calibration), and method blank analysis begins at the moment of injection of
the DFTPP or BFB analysis that the laboratory submits as documentation of
instrument performance.  The time period ends after 12 hours have elapsed
according to the system clock.  For pesticide/Aroclor analyses performed by
GC/EC, the twelve hour time period in the analytical sequence begins at the
moment of injection of the instrument blank that precedes sample analyses,
and ends after twelve hours have elapsed according to the system clock.

VALIDATED TIME OF SAMPLE RECEIPT (VTSR) - the date on which a sample is
received at the Contractor's facility, as recorded on the shipper's delivery
receipt and Sample Traffic Report.

VOLATILE COMPOUNDS - compounds amenable to analysis by the purge and trap
technique.   Used synonymously with purgeable compounds.

WIDE BORE CAPILLARY COLUMN - a gas chromatographic column with an internal
diameter (ID) that is greater than 0.32 mm.  Columns with lesser diameters
are classified as narrow bore capillaries.
                                   G-6                          OLM01.2  1/91

-------
                EXHIBIT H
  DATA DICTIONARY AND  FORMAT  FOR  DATA
DELIVERABLES IN COMPUTER-READABLE FORMAT
                                                   OLM01.0

-------
                             AGENCY STANDARD IMPLEMENTATION
                                  FOR ORGANICS OLM01.0
            (This document will replace existing formats as of October 1991)


 1.     Format  Characteristics

 1.1    This constitutes an  implementation of the EPA Agency Standard for Electronic Data
       Transmission based upon analytical results and ancillary information  required by
       the contract.  All data generated by a single analysis are grouped together, and
       the groups are aggregated to produce files that report data from an SDG.  Because
       this implementation  is only a subset of the Agency Standard, some fields have
       been replaced by delimiters as place holders for non-CLP data elements.

 1.2    This implementation  includes detailed specifications for the required format of
       each record.  The position in the record where each field is to be contained
       relevant to other fields is specified, as well as the maximum length  of the
       field.  Each field's required contents are specified as literal (contained  in
       quotes), which must  appear exactly as shown (without quotes), or as a variable
       for which format and/or descriptions are listed in the format/contents column.
       Options and examples are listed for most fields.  For fields where more than
       three options are available, a list and description of options are supplied on a
       separate page following the record descriptions.  Fields are separated from each
       other by the delimiter "|" (ASCII 124).  Fields that do not contain data should
       be zero length with  the delimiter as place holder.

 1.3    Numeric fields may contain numeric digits, a decimal place, and a leading minus
       sign.  A positive sign is assumed if no negative sign is entered in a numeric
       field and must not be entered into any numeric field.

       Requirements for significant figures and number of decimal places are specified
       in Exhibit B.  The numeric field lengths are specified such that all  possible
       numeric values can be written to the file.  The size of the numeric field
       indicates the maximum number of digits, decimal, and negative sign if appropriate
       that can appear in the field at the same time.  Therefore, the number reported
    •   may need to be rounded (using EPA Rounding Rules) to fit into the field.  The
       rounding must maintain the greatest significance possible providing the field
       length limitation.  In addition,  the rounded number that appears on the form, and
       therefore the field in the diskette file, must be used in any calculation that
      may result in other numbers reported on the same form or other forms  in the SDG.
       Field lengths should only be as long as necessary to contain the data; packing
      with blanks is not allowed.

2.     Record Types

2.1   The Agency Standard consists of variable length ASCII records.  Maximum field
      length specifications match the reporting requirements in Exhibit B.   The last
      two bytes of each record must contain "carriage return" and "line feed",
      respectively.
                                        H-l              .            OLM01.8  8/91

-------
2.2
2.3
3.
This  implementation consists of eleven record types that can be summarized in
four  groups, designated by the first record type in each group:
          Type

     Run Header
              Type ID

                10
     Sample Header    20

     Results Record  . 30
     Comments Record  90
                    Contents

Information pertinent to a group of samples processed
in a continuous sequence;  usually several per SDG

Sample identifying, qualifying,  and linking
information
Analyte results and qualifications
Free form comments
A separate run header is used for volatiles, semivolatiles,  and each column
analysis for pesticides (minimum of four Type 10 series for VOA/SV/PEST SDG).
The 20 series records are used to link samples within an SDG to the corresponding
calibrations, blanks, and so on for screening purposes.  The 30 series records
contain the actual analytical results by analyte within each sample.  The 10, 20,
and 30 records are associated with each other by their position in the file
(i.e., 30 series records follow the corresponding 20 series, which in turn follow
the 10 series run header records).

Production Runs

A production run represents a "group" or "batch" of samples that are processed in
a continuous sequence under relatively stable conditions.   Specifically:

Calibration - All samples in a run use the same initial calibration data.

Method number - Constant.
      Instrument conditions - Constant throughout a run.
      instruments cannot be combined in one run.
                                                    Results obtained on different
      Analyses from each fraction consist of separate production runs,  and are reported
      in separate files.  There will be a separate production run for each 72-hour
      sequence for pesticides for each GC column utilized.   Thus,  a full three fraction
      analysis will consist of a minimum of four production runs,  and could consist of
      more.
                                        H-2
                                                              OLM01.8   8/91

-------
  EXAMPLE OF THE SEQUENCE OF RECORD TYPES IN A FILE
10        Contains Run Header information
     II        Contains additional run-wide information if required.
     20        Occurs once for each sample,  calibration,  mean response factor,
               matrix spike duplicate result,  etc.  -  Acts as  a header.
     21
     22        Contains additional information for  samples.
     23
     27
          30        Occurs once for each final analytical result.   Reports
                    the value being determined as defined by  the type 20.
               32        Reports any auxiliary data necessary.
               33        Reports compound names for TICs  if necessary.
               36        Reports any instrumental data necessary.
          30        Values for the next  analyte or  parameter  being measured.
                         Additional data may vary for each parameter, and records
               32        may occur in any order.  Multiple occurrences of the
               33        same record type,  however, must  be consecutive.
               36
          30        Continues for as many as are necessary.
               32
               33
               36
          30
               32
               33
               36
     20        Next Sample Header record -  The following  applies to the next
     21        sample or other group of  data.
     22
          30
               32
               33
               36
          30
               32
               33
               36
                         etc.
     20
     21
          30
               32
               33
               36
                         etc.
                                    H-3                         OLM01.8  8/91

-------
 4.     Record  Sequence

 4.1    The  sequence  of  records  for Agency Standard files is as follows:  A Run Header
       (type 10)  record must be present once and once only (per file) as the first
       record  in  a file.  Therefore, a complete VOA/SV/PEST SDG will consist of several
       files.

 4.2    Each environmental sample, calibration, or quality control sample is represented
       by a group composed of type 20, 21, 22, 23, and 27 records, that hold sample
       level identifying information, followed by type 30, 32, 33, and 36 records for
       each method analyte or standard in the sample.  The type 20 record holds a count
       for  the number of method analytes being determined, and includes all target
       compounds, surrogates, internal standards plus each peak of the multi-component
       pesticides (do not include TICs in this count).   A separate field on the Type 23
       record  contains  the number of TICs found.  Type 20 records must occur in the
       order of sample  analysis.  In addition, a type 20 record is used as a header for
       any  additional run-wide data that must be reported for each method analyte (such
       as mean response factors).  Unique identifiers given in Section 10 are used in
       place of "QC  codes" to indicate the type of data that follows.  Type 30 records
       for  each analyte must occur in the order specified on hardcopy deliverable Form
       6.

 4.3    Type 90 comment  records may be defined to occupy any position except before the
     .  type 10 (header) record.

 5.     File/Record Integrity

       All  record types shall contain the following check fields to ensure file and
       record integrity:
     Record
     Position

    First Field
             Field
             Length
Field
Contents

Record type
Remarks

"10" or as appropriate
    Last Field
                    4
                    2
                       Record sequence number
                       within file
                       Record checksum
                       Must contain CR and LF
                               00000-99999, numbered
                               sequentially
                               Four hexadecimal digits(*)
6.
(*)  The checksum is the sum of the ASCII representation of the data on the
record up to the Record Sequence Number plus the checksum of the previous record.
The sum is taken modulo 65536 (2  ) and represented as four hexadecimal digits.

Dates and Times
      Date or time-of-day information consists of successive groups of two decimal
      digits, each separated by delimiters.  Dates are given in the order YY MM DD, and
      times as HH MM.  All hours must be given as 00 to 23 using a 24 hour-clock and
      must be local time.
                                        H-4
                                                               OLM01.8   8/91

-------
7.    Multiple Volume Data

      There is no requirement under this format that all the data from an entire sample
      delivery group fit onto a single diskette.  However,  each single production run
      must fit onto a single diskette if possible.  If that is not possible, then it is
      necessary that all files start with a type 10 record, and that the multiple type
      10 records for each file of the same production run be identical.  Information
      for a single sample may not be split between files.

8.    Deliverable

8.1   The file must be submitted on 5-1/4 inch floppy diskette(s),  which may be either
      double-sided, double density, 360 K-byte or  high capacity 1.2 M-byte
      diskette(s).  IBM-compatible, 3.5 inch double-sided,  double density 720 K-byte or
      high density 1.44 M-byte diskettes may also be submitted.  The diskettes must be
      formatted and recorded using MS-DOS Operating System.  The diskettes must contain
      all information relevant to one and only one SDG, and must accompany the hardcopy
      package for the SDG submitted to the Sample Management Office (see Exhibit B) .

      Agency Standard data from an entire SDG may not fit onto a single diskette.   If a
      single production run is being split onto multiple diskettes, then all files  must
      start with a type 10 record, and the multiple type 10 records for each file of
      the same production run must be identical.  Do not split the data from a single
      sample onto multiple diskettes.

8.2   Information on the diskette must correspond to information submitted in the
      hardcopy raw data package and on the hardcopy raw data package forms.   For
      example, type 30 results field specifies maximum length of 13.  When reporting
      CRQLs or results on Form 1 maximum length is 13 as is specified on Exhibit H;
      when reporting 'calculated amounts' on Form 7D, hardcopy specified maximum length
      of 8 and so, a maximum length of 8 should be used.  Blank or. unused records must
      not be included on the diskettes.   If the information submitted in the hardcopy
      data package forms is changed, the information in the diskette file must be
      changed accordingly,  and a complete diskette containing all the information for
      the SDG must be resubmitted along with the hard copy at no additional cost to the
      EPA.

8.3   Each diskette must be identified with an external label containing (in this
      order)  the following information:

               Disk Density
               File Name(s)
               Laboratory Name (optional)
               Laboratory Code
               Case Number
               SAS Number (where applicable)

      The format for the File Name must be XXXXX.001 to XXXXX.099
               where  XXXXX   is the SDG identifier, 0 designates organics,  and 01
                              through 99 the file number.

      Dimensions of the label must be in the range 4-3/4" to 5" long by 1-1/4" to 1-
      1/2"  wide for 5-1/4 inch floppy diskette; and 2" to 2-1/4" long by 2-1/8" to  2-
      3/8"  wide for 3.5 inch IBM-compatible diskette.

                                        H-5                         OLM01.8   8/91

-------
9 .    Record Listing

      Following  is a  listing of every record type required to report data from a single
      SDG.
PRODUCTION RUN HEADER RECORD (Type 10)

MAXIMUM
LENGTH    CONTENTS

2         RECORD TYPE
6         Delimiters
5         MEASUREMENT TYPE
1         Delimiter
8         METHOD NUMBER
2         Delimiters
6         LAB CODE
4         Delimiters
11        CONTRACT NUMBER
1         Delimiter
10        INSTRUMENT ID
2         Delimiters
25        LABORATORY NAME
2         Delimiters
5         RECORD SEQUENCE NUMBER
4         CHECKSUM
                                                      FORMAT/
                                                      CONTENTS

                                                      "10"
                                                      I I I I I I
                                                      Character
                                                      |
                                                      Character
                                                      | |
                                                      Character
                                                      | | | |
                                                      Character
                                                      |
                                                      Character
                                                      | |
                                                      Character
                                                      | |
                                                      Numeric
                                                      Character
1   General  descriptor GC/MS  or GC.

2   OLM01.0V For Volatiles; OLM01.0B  For  Semivolatiles ;  OLM01.0P For .Pesticides .
                                   H-6
                                                                    OLM01.8   8/91

-------
CHROMATOGRAPHY RECORD (TYPE 11)

  Use:  To describe chromatograph condition.  Must be present for volatiles and
        pesticides.  Is optional for semivolatiles.
MAXIMUM                                          FORMAT/
LENGTH    CONTENTS                               CONTENTS

2         RECORD TYPE                            "11"
1         Delimiter                              |
10        GC COLUMN                              Character
2         Delimiters                             ||
4         GC COLUMN ID                           Numeric (mm)
11        Delimiters                             I I I II I I I I I I
5         RECORD SEQUENCE NO.                     Numeric
4         CHECKSUM                               Character
                                   H-7                         OLM01.8   8/91

-------
     SAMPLE HEADER DATA RECORD (TYPE 20)
     MAXIMUM
     LENGTH    CONTENTS

     2         RECORD TYPE
     2         Delimiters
     12        EPA SAMPLE NUMBER
     1         Delimiter
     1         MATRIX
     1         Delimiter
     3         QC CODE
     1         Delimiter
     3         SAMPLE QUALIFIER
     1         Delimiter
     5         CASE NUMBER
     1         Delimiter
     6         SDG NO.
     1         Delimiter
     2         YEAR ANALYZED
     1         Delimiter
     2         MONTH ANALYZED
     1         Delimiter
     2         DAY ANALYZED
     1         Delimiter
     2         HOUR ANALYZED
     1         Delimiter
     2         MINUTE ANALYZED
     2         Delimiters
     2         SAMPLE WT/VOL UNITS
     1         Delimiter
     5         SAMPLE WT/VOL
     1         Delimiter
     3         ANALYTE COUNT
     3         Delimiters
     5         RECORD SEQUENCE NO.
     4         CHECKSUM
FORMAT/
CONTENTS

"20"
II
As is exactly on
the hardcopy form

CHARACTER1
I
Character (See Section 10)

1               2
RIN/REX/REJ/SRN^
I
Numeric
I
Character
I
YY
I
MM
I
DD
I
HH
I
MM
II
"G"/"ML"
Numeric
I       ,
Numeric
III
Numeric
Character
1   "0"  if not applicable (calibration,  tune,  etc.),"l"  for water,  "H"  for soil.

2   "RIN"  for reinjection,  "REX"  for re-extractions,  "REJ"  for  rejected samples,  and "SRN"
    for  dilutions.

3   Sample WT/VOL is the volume  in mililiters  for  liquid and the  wet weight in grams for
    solids.   The sample units  code indicates which units are in use for the current
    sample.   Leave  zero or  blank if not  applicable.   Sample WT/VOL includes purge volume.

4   1-3  decimal digits;  Counts  all analytes including surrogates,  internal standards,  and
    all  peaks for multi-component pesticides.   For calibrations,  count  also DFTPP/BFB if
    mixed  in injection.
                                        H-8
               OLM01.8  8/91

-------
      SAMPLE  HEADER DATA RECORD  (TYPE  21)
        Use:   Continuation of Type  20
        Position:   Follows  the Type  20  to which it applies.
     MAXIMUM
     LENGTH     CONTENTS
      2         RECORD TYPE
      1         Delimiter
      1         PURGE
      1         Delimiter
      1         LEVEL
      2         Delimiters
      1         EXTRACTION
      2         Delimeters
      6         SAS NUMBER
      1         Delimiter
      14        LAB FILE/SAMPLE ID
      1         Delimiter
      2         YEAR EXTRACTED
      1         Delimiter
      2         MONTH EXTRACTED
      1         Delimiter
      2         DAY EXTRACTED
      2         Delimiters
      2         YEAR RECEIVED
      1         Delimiter
      2         MONTH RECEIVED
      1         Delimiter
      2         DAY RECEIVED
      2         Delimiters
      8         INJECTION/ALIQUOT VOLUME
      2         Delimiters
      5         RECORD SEQUENCE NO.
      4         CHECKSUM
FORMAT/
CONTENTS

"21"
I
"N" for not heated; "Y" for heated
II
Character

1
Character
I
YY
I
MM
DD
II
YY
I
MM
I
DD
II     ,
Numeric
II
Numeric
Character
1   "L"  for low level samples and "M" for medium level samples for volatile and
    semivolatile analyses.   Leave blank for calibrations and tunes.

2   "S"  for separatory funnel,  "C" for continuous liq-liq and "N" for sonication.

3   Lab  file ID for Volatile, Semivolatile; Lab sample ID for Pesticides in same format as
    on forms.

4   Injection volume for BNAs and PESTs;  Soil Aliquot Volume for VOA.
                                         H-9
               OLM01.8   8/91

-------
     SAMPLE CONDITION RECORD (TYPE 22)

      Use:  Continuation of Type 20.  Used to describe additional Sample Conditions

      Position:   Follows the Type 20 and 21 to which it applies.
MAXIMUM
LENGTH
2
1
2
1
2
1
2
1
2
1
2
1
14
1
4
1
5
1
1
1
8
1
8
3
5
1
5
4
1 For average
2 If "AVERAGE

CONTENTS
RECORD TYPE
Delimiter
CALIBRATION YEAR
Delimiter
CALIBRATION MONTH
Delimiter
CALIBRATION DAY
Delimiter
CALIBRATION HOUR
Delimiter
CALIBRATION MINUTE
Delimiter
CALIBRATION FILE ID
Delimiter
PH
Delimiter
PERCENT MOISTURE
Delimeter
DECANTED
Delimeter
EXTRACT VOLUME
Delimiter
DILUTION FACTOR
Delimiters
LEVEL
Delimiter
RECORD SEQUENCE NO.
CHECKSUM
, use the date and time
" is entered, then Form
FORMAT/
CONTENTS
II O O II
1
YY1
1
MM
1
DD
1
HH
1
MM
1
2
Character
1
Numeric
1
Numeric
1
"Y" or "N"
1
3
Numeric
|
1 4
Numeric
Ml
5
Numeric
1
Numeric
Character
average was calculated.
8 will be constructed usii
    with a QC code of "CLD".   This field must match Lab  Field/Sample  ID on Type  21  for  the
    associated calibration.

3   Use  the initial extract  volume adjusted (multiplied)  by all  contract-mandated
    dilutions that are to be excluded from the dilution  factor.   Soil extract volume for
    VOA;  Cone.  Extract volume for BNA and Pest.   (In microliters)

4   Dilution factor of sample analyzed (omit contract-mandated dilutions).

5   Concentration level of Pesticide Individual  Mix A and B standards.   Concentration of
    low  point,  mid point and high point calibration standards as a multiplier of low
    point.   Low point =1.0,  Mid point = up to 10.0,  High point  = 30.0 to  100.0.


                                        H-10                         OLM01.8   8/91

-------
ASSOCIATED INJECTION AND COUNTER RECORD (TYPE 23)

  Use:  Continuation of Type 20.  Used to identify associated blanks and  tunes,  and
        to count  the number of surrogates and spikes outside of the QC  limits  and
        the number of TIC compounds.  Used for Forms 3,4, and 5.

  Position:  Follows the. Type 20, 21, and 22 to which it applies.
MAXIMUM
LENGTH    CONTENTS

2         RECORD TYPE
1         Delimiter
1         TUNE LABEL
1         Delimiter
2         INJECTION YEAR
1         Delimiter
2         INJECTION MONTH
1         Delimiter
2         INJECTION DAY
1         Delimiter
2         INJECTION HOUR
1         Delimiter
2         INJECTION MINUTE
1         Delimiter
14        DFTPP/BFB LAB FILE ID
1         Delimiter
2         INSTRUMENT BLANK LABEL
1         Delimiter
2         BLANK INJECTION YEAR
1         Delimiter
2         BLANK INJECTION MONTH
1         Delimiter
2         BLANK INJECTION DAY
1         Delimiter
2         BLANK INJECTION HOUR
1         Delimiter
2         BLANK INJECTION MINUTE
1         Delimiter
14        INSTRUMENT BLANK LAB FILE/SAMPLE ID
4         Delimiters
2         METHOD BLANK LABEL
1         Delimiter
2         BLANK INJECTION YEAR
1         Delimiter
2         BLANK INJECTION MONTH
1         Delimiter
2         BLANK INJECTION DAY
1         Delimiter
2         BLANK INJECTION HOUR
1         Delimiter
2         BLANK INJECTION MINUTES
1         Delimiter
14        METHOD BLANK LAB FILE/SAMPLE ID
1         Delimiter
FORMAT/
CONTENTS

 "23"
  I
 "P" or blank
  I
  YY
  I
  MM
  I
  DD
  I
  HH
  I
  MM
  I
  Character
  I
 "IB" or blank
  I
  YY
  I
  MM
  I
  DD
  I
  HH
  I
  MM
  I
  Character
  INI
 "MB"
  I
  YY
  I
  MM
  I
  DD
  I
  HH
  I
  MM
  I
  CHARACTER
                                   H-ll
               OLM01.8  8/91

-------
     ASSOCIATED INJECTION AND COUNTER RECORD (TYPE 23) CONT:

     MAXIMUM                                          FORMAT/
     LENGTH    CONTENTS                               CONTENTS

    1          SURROGATE RECOVERY LABEL              "P" for % recoveries
    1          Delimiter                              |
    2          SURROGATE RECOVERIES OUT               Numeric
    1          Delimiter                              |
    1          TIC LABEL                              T" for TICS
    1          Delimiter                              |
    2          NO. OF TICS                            Numeric
    1          Delimiter                              |
    1          SPIKE RECOVERY LABEL                   "S" for Spikes
    1          Delimiter                              |
    2          SPIKE RECOVERIES OUT                   Numeric
    1          Delimiter                              |
    1          RPD LABEL                              "R" for RPD1
    1          Delimiter                              |
    2          RPD OUT                                Numeric
    1          Delimiter                              |
    5          RECORD SEQUENCE NO.                    Numeric
    4          CHECKSUM                               Character
1   "R"  for  Matrix Spike/Duplicate  Recovery  Relative  Percent Differences.


                                        H-12                         OLM01.8   8/91

-------
     SAMPLE CLEANUP RECORD (TYPE 27)
        Use:   Continuation  of Type  20.  Used to idencify sample/blank cleanup procedures
              and QC  results.  Used for Form 9.
        Position:  Follows  type 20,  21, 22, and 23 to which it applies
     MAXIMUM
     LENGTH    CONTENTS

     2         RECORD TYPE
     1         Delimiter
     1         FIRST CLEAN-UP TYPE
     1         Delimiter
     2         GPC CALIBRATION CHECK YEAR
     1         Delimiter
     2         GPC CALIBRATION CHECK MONTH
     1         Delimiter
     2         GPC CALIBRATION CHECK DAY
     1         Delimiter
     2         GPC CALIBRATION CHECK HOUR
     1         Delimiter
     2         GPC CALIBRATION CHECK MINUTE
     1         Delimiter
    14         GPC Data Descriptor
     1         Delimiter
     1         SECOND CLEAN-UP TYPE
     1         Delimiter
     2         FLORISIL LOT CHECK YEAR
     1         Delimiter
     2         FLORISIL LOT CHECK MONTH
     1         Delimiter
     2         FLORISIL LOT CHECK DAY
     1         Delimiter
     2         FLORISIL LOT CHECK HOUR
     1         Delimiter
     2         FLORISIL LOT CHECK MINUTE
     1         Delimiter
     14        FLORISIL DATA DESCRIPTOR
     1         Delimiter
     1         SULFUR CLEAN-UP
     1         Delimiter
     2         SULFUR BLANK LABEL
     1         Delimiter
     2         BLANK INJECTION YEAR
     1         Delimiter
     2         BLANK INJECTION MONTH
     1         Delimiter
FORMAT/
CONTENTS

"27"
I
"G" for GPC
I
YY
I
MM
I
DD
I
HH
I
MM
I         ,
Character
I
"F" or blank
I
YY
I
MM
I
DD
I
HH
I
MM

1         2
Character
I
Y/N
I
"SB"
I
YY
I
MM
1   Lab  Sample ID for GPC (GPC Column).   Format is "GPC"  followed by unique identifier.

2   Lab  Sample ID for Florisil lot check.   Format is "FLO"  followed by florisil cartridge
    lot  number.
                                        H-13
               OLM01.8  8/91

-------
 SAMPLE CLEANUP RECORD (TYPE 27) Cont.

 MAXIMUM                                          FORMAT/
 LENGTH    CONTENTS                               CONTENTS

 1         Delimiter                              |
 2         BLANK INJECTION DAY                    DD
 1         Delimiter                              I
 2         BLANK INJECTION HOUR                   HH
 1         Delimiter                              |
 2         BLANK INJECTION MINUTE                 MM
 1         Delimiter                              |
14         SULFUR BLANK LABORATORY FILE/SAMPLE ID Character
 1         Delimiter                              |
 5         RECORD SEQUENCE NO.              .      Numeric
 4         CHECKSUM                               Character
                                   H-14                         OLM01.8  8/91

-------
RESULTS DATA RECORD (TYPE 30)
MAXIMUM
LENGTH
CONTENTS
FORMAT/
CONTENTS
2
1
1
1
9
1
9
1
5
1
3
1
13
1
5
I
I
1
13
1
1

1
13
1
1
1
5
1
1
1
5
1
5
1
1
1
5
1
1
1
5
1
1
1
13
1
1
1
RECORD TYPE
Delimiter
ANALYTE LABEL
Delimiter
CAS NUMBER
Delimiter
INTERNAL STD. CAS NUMBER
Delimiter
CONCENTRATION UNITS
Delimiter
RESULT QUALIFIER
Delimiter
RESULTS
Delimiter
FLAGS
Delimiter
AMOUNT ADDED LABEL
Delimiter
AMOUNT ADDED
Delimiter
CRQL LABEL

Delimiter
CRQL
Delimiter
RSD LABEL
Delimiter
RSD VALUE .
Delimiter
MS/MSD REC LABEL
Delimiter
MS % RECOVERY
Delimiter
MSD % RECOVERY
Delimiter
RPD LABEL
Delimiter
RPD VALUE
Delimiter
SURR/SPIKE RECOVERY LABEL
Delimiter
SURR/SPIKE RECOVERY
Delimiter
MEAN CONCENTRATION LABEL
Delimiter
MEAN CONCENTRATION
Delimiter
PERCENT DIFFERENCE LABEL
Delimiter
"30"
I
"C" for Cas Number
I
Numeric
I
Numeric
Character
I         !
Character
I
Numeric

1         2
Character

"A" for Amt. added3
I
Numeric
I
"U" for "undetected"
 or blank
I
Numeric
I
"R" for RSD

I       4
Numeric
I
"P" for % recovery
I
Numeric
I
Numeric
I
"D"
i       5
Numeric
I
"S" for % recovery
I          ,
% Recovery
I
"M" for Mean cone.
I       7
Numeric

1          8
"F" or "P"°
                                   H-15
                                                      OLM01.8  8/91

-------
     RESULTS DATA RECORD (TYPE 30) CONT:

     5         PERCENT DIFFERENCE                     Numeric
     1         Delimiter                              |
     1         INTERNAL STANDARD AREA LABEL           "I" for IS Area
     1         Delimiter                              |
     13        INTERNAL STANDARD AREA                 Numeric
     1         Delimiter                              |
     5         RECORD SEQUENCE NO.                    Numeric
     4         CHECKSUM                               Character

1   For Type 20 for calibration summary  (MNC),  use  "AVG"  for  average RRFs  and Mean
    Calibration Factors.

2   A maximum of five flags (D,E,J,B,A,P or  N)  with no space  between the flags can be
    reported,  each representing a qualification of  the result as  described in Exhibit B.

3   Also Nominal Amount  for Pest  Form (7D-7E).

4   "R" for % Resolution (Form 6G)  or for RSD of Response factors under Calibration
    summary (MNC)  Type 20.

5   RPD for MS/MSD recoveries,  or for Pest.  Calibration Verification (Form 7D/7E).

6   Surrogate or Spike (Forms  2,  Form 9A/9B) recovery.

7   Mean Concentration for  Multicomponent analytes  detected in Pesticide analyses.

8   "P" for Percent  Difference between concentrations from two columns  in  Pesticide
    analyses,  or "F" for  Percent  Difference  between average RRF (initial calibration) and
    RRF50 (continuing calibration)  in VOA/BNA analyses.
                                        H-16                         OLM01.8   8/91

-------
  AUXILIARY DATA RECORD (TYPE 32)

    Use:  Used to report scan number and retention time (in minutes) for Internal
          Standards and for TIC compounds.   Used to report retention time data and
          Percent Breakdown for Pesticides.
    Position:  Follows Type 30.   (Record will  only be  required as specified above.)
  MAXIMUM
  LENGTH    CONTENTS

  2         RECORD TYPE
  3         Delimiters
  2         RETENTION TIME LABEL
  1         Delimiter
  5         RETENTION TIME
  1         Delimiter
  3         FIRST LIMIT LABEL
  1         Delimiter
  5         RT WINDOW LOWER LIMIT
  1         Delimiter
  3         SECOND LIMIT LABEL
  1         Delimiter
  1         RT WINDOW UPPER LIMIT
  2         Delimiters
  2         THIRD LIMIT LABEL
  1         Delimiter
  5         % BREAKDOWN
  1         Delimiter
  5         COMBINED % BREAKDOWN
  2         Delimiters
  1         PEAK
  1         Delimiter
  5         RECORD SEQUENCE NO.
  4         CHECKSUM
FORMAT/
CONTENTS

"32"
III
"RT"
I
Numeric
I
"RTF"
I
Numeric
I
"RTT"
I
Numeric
II
"PB" for % breakdown
I
Numeric (DDT/ENDRIN)
I
Numeric

11         1
1 THROUGH 51
I
Numeric
Character
Peaks 1, 2, and 3 are mandatory, peaks 4 and 5 are optional.  Types  30  and  31  will  be
repeated for each peak that is reported (a minimum of three, a maximum  of five times).
This is for multicomponent analytes in PESTICIDE analyses.
                                    H-17
               OLM01.8  8/91

-------
NAME RECORD (Type 33)

      Use:  To carry an analyte name for TIC compounds

      Position:  Follows Type 30 for TIC compounds.

    MAXIMUM                                          FORMAT/
    LENGTH    CONTENTS      •                         CONTENTS

    2         RECORD TYPE                            "33"
    1         Delimiter                              |
    67        NAME OF COMPOUND                       Character
    1         Delimiter                              |
    5         RECORD SEQUENCE NO.                     Numeric
    4         CHECKSUM                               Character
                                       H-18                         OLM01.8   8/91

-------
INSTRUMENTAL DATA READOUT RECORD (TYPE 36)
  Use:  To describe DFTPP/BFB percent abundances for Form 5.

  Position:  Follows Type 30 for internal standards and DFTPP/BFB data.
MAXIMUM
LENGTH    CONTENTS

2         RECORD TYPE
1         Delimiter
1         MASS LABEL
3         Delimiters
3         FIRST MASS (DFTPP/BFB)
2         Delimiters
5         FIRST PERCENT ABUNDANCE
1         Delimiter
3         SECOND MASS
1         Delimiter
5         SECOND PERCENT ABUNDANCE
1         Delimiter
5         PERCENT MASS
1         Delimiter
3         THIRD MASS
1         Delimiter
5         THIRD PERCENT ABUNDANCE
2         Delimiters
3         FOURTH MASS
1         Delimiter
5         FOURTH PERCENT ABUNDANCE
1         Delimiter
5         PERCENT MASS
1         Delimiter
3         FIFTH MASS
1         Delimiter
5         FIFTH PERCENT ABUNDANCE
1         Delimiter
5         PERCENT MASS
1         Delimiter
3         SIXTH MASS
1         Delimiter
5         SIXTH PERCENT ABUNDANCE
2         Delimiters
3         SEVENTH MASS
1         Delimiter
5         SEVENTH PERCENT ABUNDANCE
1         Delimiter
5         PERCENT MASS
1         Delimiter
3         EIGHTH MASS
1         Delimiter
5         EIGHTH PERCENT ABUNDANCE
1         Delimiter
5         PERCENT MASS
FORMAT/
CONTENTS

"36"
I
"M"
Ml
Numeric
II
Numeric
I
Numeric
I
Numeric
I
% mass of 69, BNA only
I
Numeric
I
Numeric
II
Numeric
I
Numeric
I
% mass of 69, BNA only
I
Numeric
I
Numeric
I
% mass of 174, VOA only
I
Numeric
I
Numeric
II
Numeric
I
Numeric
I
% mass of 174, VOA only
I
Numeric
I
Numeric
I
% mass of 174, VOA only
                                   H-19
               OLM01.8  8/91

-------
INSTRUMENTAL DATA READOUT RECORD (TYPE 36) CONT:
MAXIMUM
LENGTH    CONTENTS

1         Delimiter
3         NINTH MASS
1         Delimiter
5         NINTH PERCENT ABUNDANCE
1         Delimiter
5         PERCENT MASS
1         Delimiter
3         TENTH MASS
1         Delimiter
5         TENTH PERCENT ABUNDANCE
2         Delimiters
3         ELEVENTH MASS
1         Delimiter
5         ELEVENTH PERCENT ABUNDANCE
2         Delimiters
3         TWELFTH MASS
1         Delimiter
5         TWELFTH PERCENT ABUNDANCE
2         Delimiters
3         THIRTEENTH MASS
2         Delimiters
5         THIRTEENTH PERCENT ABUNDANCE
1         Delimiter
5         PERCENT MASS
1         Delimiter
5         RECORD SEQUENCE NO.
4         CHECKSUM
FORMAT/
CONTENTS

I
Numeric
I
Numeric
I
% mass of 176, VOA only
I
Numeric
I
Numeric
II
Numeric
I
Numeric
II
Numeric
I
Numeric
II
Numeric
II
Numeric
I
% mass of 442, BNA only
I
Numeric
Character
                                   H-20
               OLM01.8  8/91

-------
COMMENT RECORD (Type 90)

       Use:  To provide for Operator-Entered Comments.

       Position:  May occur anywhere

     MAXIMUM                                          FORMAT/
     LENGTH    CONTENTS                               CONTENTS

     2         RECORD TYPE                            "90"
     1         Delimiter                              |
     67        ANY COMMENT                            Character
     1         Delimiter                              |
     5         RECORD SEQUENCE NO.                     Numeric
     4         CHECKSUM                               Character
                                       H-21                         OLM01.8  8/91

-------
10.    Definitions of Various  Codes  Used  in Agency Standard Records

10.1  Quality Control and Related Codes  (QCC)  in Type 20 Records

      Note:   These codes appear in  the QC code fields of type  20  records.   They are  used
             to indicate the  type of data that is being reported.
    QCC
Name
    LRB   LABORATORY (REAGENT)
          BLANK

    LIB   LABORATORY INSTRUMENT
          BLANK

    LSB   LABORATORY SULFUR
          BLANK
                Definition

The "Method Blank"  (See Exhibit G).


The "Instrument Blank"


If different from "Method Blank"  (Pesticides)
    LSD    LABORATORY SPIKE
          DUPLICATE  BACKGROUND
          (ORIGINAL)  VALUES
    LF1    LABORATORY  SPIKED
          SAMPLE  -  FINAL  -
          FIRST MEMBER

    LF2    LABORATORY  SPIKED
          SAMPLE  -  FINAL  -
          SECOND  MEMBER
                        An environmental  sample  which is analyzed according
                        to the analytical method,  and subsequently used for
                        the matrix spike  and the matrix spike duplicate
                        (See Exhibit G).

                        The "Matrix Spike"  (See Exhibit G)  - must preceed
                        LF2
                        The "Matrix Spike  Duplicate"   (See Exhibit G)
   LPC   LABORATORY PERFORMANCE
         CHECK SOLUTION
                        A solution of DFTPP  (BNA)  or BFB (VGA)  or
                        method analytes  (PEST/PCB)  used to evaluate the
                        performance of an instrument with respect to a
                        defined set of criteria (Tune or Resolution Check
                        Sample)  (See Exhibit  G).
   FLO   FLORISIL CHECK
         SOLUTION

   GPC   GPC CHECK
         SOLUTION
                        A solution of pesticides  used to check recovery
                        from each lot of Florisil cartridges.

                        A solution of pesticides  used to check recovery
                        from each new GPC calibration.
   CLM   INITIAL CALIBRATION  -
         MULTI POINT
                        The  Initial  Calibration for  GC/MS (See Exhibit G),
                        or the  Initial  Individual  Standard Mixes (A,
                        B) for  Pesticides  (See  Exhibit D PEST).
                        Response  factors  (GC/MS) or  Calibration Factors
                        (Pesticides)  rather  than concentrations will be
                        reported  on  the following  type 30 records.
                                       H-22
                                                        OLM01.8  8/91

-------
    CLS   INITIAL CALIBRATION       The Initial Toxaphene/Aroclor Mixes used to determine
          SINGLE POINT              all calibration factors.  (See Exhibit D PEST).

    CLC   CONTINUING CHECK          The continuing calibration for GC/MS (See Exhibit G).
          CALIBRATION

    CLE   CONTINUING PERFORMANCE    The subsequent Individual Standard Mixes
          CHECK                     (A,B) and Performance Evaluation Mixture for
                                    Pesticides (See Exhibit D PEST).

    CLD   DUAL PURPOSE              A calibration solution as above used both as an
          CALIBRATION               initial calibration (CLM) and a continuing check
                                    (CLC).  [50 level initial calibration if needed
                                    for Form 8]


    blank                          Sample, not associated with any quality control
                                   item.

      The following QCC values are used on type 20 records which act as a header, and
      indicate that additional (usually calculated) analyte specific data will be present
      on type 30 (and following type) records.  Usually these data will apply to an
      entire production run, in which case they will appear immediately following the
      type 10 record.  If the data apply to only a portion of the samples in the run,
      they should be placed immediately preceding the samples to which they apply.  Much
      of the rest of the information in the type 20 record may be blank, indicating that
      these data do not apply to these results.

     MNC   MEAN VALUES FROM          The data following represent mean values and
           CALIBRATIONS              percent RSD's from the initial calibration
                                     (GC/MS) or the mean calibration factors, mean
                                     retention times amd retention time windows
(pesticides).

10.2  Codes For Sample Medium (Matrix1, Source)

           Medium                                                    Code

           All Media, Specific Medium not Applicable.  Use for         0
           Calibrations,  Tunes,  etc

           Water                                                       1
                                        H-23                         OLM01.8   8/91

-------
10.3  List of Sample and Result Qualifiers
     Definition:   A sample qualifier (also called a non-numeric  result)  consists  of 3
                  characters which act as an indicator of the  fact and the reason that
                  the subject analysis (a) did not produce a numeric  result,  or (b)
                  produced a numeric result for an entire sample but  it  is qualified in
                  some respect relating to the type or validity  of the result.
      10.3.1  Sample Qualifiers

       Qualifier    Full Name
        RIN
        REX
        REJ
 RE-ANALYZED
 RE-PREPARED
 REJECTED
        SRN
        SRS
 DILUTED
SECONDARY
DILUTION
        Definition

The indicated analysis results were
generated from a re-injection of
the same sample extract or aliquot

The indicated analysis results were
generated from a re-extraction of
the same sample

The results for the entire sample analysis have
been rejected for an unspecified reason by the
laboratory.  For initial calibration
data,  these data were not utilized in the
calculation of the mean.

The indicated analysis results were
generated from a dilution of the
same sample (DL SUFFIX)

The indicated analysis results were
generated from a secondary dilution of
the same sample (DL2 SUFFIX - Pesticides)
      10.3.2   Result Qualifiers in Type 30 Records

              A result qualifier (also called a non-numeric result) consists of 3
              characters which act as an indicator of the fact and the reason that the
              subject analysis (a) did not produce a numeric result,  or  (b) produced a
              numeric result for a single analyte but it is qualified in some respect
              relating to the type or validity of the result.
          BDL     BELOW  DETECTABLE  LIMITS
          NAR     NO ANALYSIS RESULT
          AVG     AVERAGE VALUE
                           Indicates compound was analyzed for but
                           not detected;   (Form 1 "U" Flag).

                           There is no analysis result required for
                           this subject parameter.

                           Average value -  used to report a range of
                           values; e.g.,  relative response factors.
                                       H-24
                                                  OLM01.8  8/91

-------
 CBC
 CANNOT BE CALCULATED
 LTL
GTL
LLS
TIE
REJ

STD



STB
LESS THAN LOWER
CALIBRATION LIMIT
GREATER THAN UPPER
CALIBRATION LIMIT
LESS THAN LOWER STANDARD
TENTATIVELY IDENTIFIED
-ESTIMATED VALUE
REJECTED

INTERNAL STANDARD
INTERNAL STANDARD
BELOW DETECTION LIMITS
The analysis result cannot be calculated
because an operand value is qualified.
Identifies analytes whose Internal
Standard is not found.

Actual value is known to be less  than  the
lower calibration range due to dilution.
(Form 1 "D" Flag)

Actual value is known to be greater than
the upper calibration range.
(Form 1 "E" Flag)

The analysis result is less than  the
sample quantitation limit.
(Form 1 "J" Flag)

The indicated analyte is a tentatively
identified analyte; its concentration has
been estimated.  (Form 1-E or 1-F "J"
Flag)

Results rejected by the laboratory.

The indicated compound is an internal
standard.  There is no analysis result  to
report.

A combination of "STD" and "BDL".
FBK
MSP
TFB
ALC
NRP
FOUND IN BLANK
PERCENT RECOVERY
 TENTATIVELY IDENTIFIED
 AND FOUND IN BLANK

 ALDOL CONDENSATION
 NON-REPRODUCIBLE
PRE
 PRESUMPTIVE PRESENCE
The indicated compound was found in the
associated method blank (LRB) as well as
the sample.  (Form 1 "B" flag)

The following value represents the percent
recovery for the "MS" sample.  .The
remaining two values give the "MSD"
percent recovery and the Percent RPD.

A Combination of "TIE" and "FBK"  (Form 1-
E or 1-F "B" flag).

Labels a suspected Aldol Condensation-
product for TIC's  (Form 1-E or 1-F "A"
Flag).

Results of two or more injections are
not comparable (Form ID "p" flag), e.g.,
Aroclor target analyte with greater than 25%
difference between column analyses.

Presumptive evidence of presence of material
for TIC (Form 1-E or 1-F "N" flag).
                             H-25
                                                  OLM01.8  8/91

-------
11.    Format of Records for Specific Uses

11.1  Format of the SAMPLE HEADER DATA RECORD  (Type 20) for Mean Response Factors
     MAXIMUM
     LENGTH    CONTENTS
     2
     3
     1
     1
     3
     2
     5
     1
     6
     9
     3
     3
     5
     4
 "20"
 Delimiters
 "0"
 Delimiter
 "MNC"
 Delimiters
 Case Number
 Delimiter
 SDG No.
 Delimiters
 Analyte  Count
 Delimiters
 Record Sequence No.
 Checksum
FORMAT/
CONTENTS

Record Type
III
All matrices
I
Identifies Mean Response Factors
II
Numeric
I
Character
MINIMI
Numeric
III
Numeric
Character
11.2  Format of the RESULTS  DATA RECORD  (Type  30) for Mean Relative  Response Factors
     MAXIMUM
     LENGTH    CONTENTS
     2
     1
     1
     1
     9
     1
     9

     3
     3
     1
     6
     6
     1
     1
     5
     14
     5
     4
"30"
 Delimiter
 "C"
 Delimiter
 CAS NUMBER
 Delimiter
 INTERNAL STANDARD
 CAS NUMBER
 Delimiters
 "AVG"
 Delimiter
 MEAN RESPONSE FACTOR
 Delimiters
 "R"
 Delimiter
 PERCENT RSD
 Delimiters
 RECORD SEQUENCE NO.
 CHECKSUM
FORMAT/
CONTENTS

Record Type
I
"C" for CAS Number
I
Numeric
I
Numeric
Ill
Indicates
I
Right Justified
Illlll
Indicates Percent
I
Numeric
IIIIIMIIIIIII
Numeric
Character
          Average Value
                  RSD
                                       H-26
                                                      OLM01.8  8/91

-------
 11.3   Format  of  the  SAMPLE HEADER DATA RECORD (Type 20) for Matrix Spike Duplicates
MAXIMUM
LENGTH

2
2
12
1
1
1
3
2
5
1
6
1
2
1
2
1
2
1
2
1
2
2
1
1
5
1
3
3
5
4
CONTENTS

Record Type
Delimiters
EPA Sample Number
Delimiter
Matrix
Delimiter
QC Code
Delimiters
Case Number
Delimiter
SDG No.
Delimiter
Year Analyzed
Delimiter
Month Analyzed
Delimiter
Day Analyzed
Delimiter
Hour Analyzed
Delimiter
Minute Analyzed
Delimiters
Sample wt/vol unit
Delimiter
Sample wt/vol
Delimiter
Analyte Count
Delimiters
Record Sequence Number
Checksum
 FORMAT/CONTENTS

"20"
 II
 As  is exactly on the Form
 I
"H"
 I
 LF2
 II
 Numeric
 I
 Character
 I
 YY
 I
 MM
 I
 DD
 I
 HH
 I
 MM
 II
"G"
 I
 Numeric
 I
 Numeric
 III
 Numeric
 Character
11.4  Format of the COUNTER RECORD (Type 23) for Matrix Spike Duplicates (for VOA and
      BNA)
      Position:  Follows the type 20 to which it applies.
MAXIMUM
LENGTH

2
2
2
1
2
1
2
1
2
1
CONTENTS

Record Type
Delimiters
Injection Year
Delimiter
Injection Month
Delimiter
Injection Day
Delimiter
Injection Hours
Delimiter
 FORMAT/CONTENTS

 "23"
 II
 YY
 I
 MM
 I
 DD
 I
 HH
 I
                                        H-27
                                                     OLM01.8   8/91

-------
2
1
14
11
2
1
2
1
2
1
2
1
2
1
2
1
14
1
1
1
2
3
1
1
2
1
1
1
2
1
5
4
Injection Minutes
Delimiter
DFTPF Lab File ID
Delimiters
Method Blank Label
Delimiter
Blank Injection Year
Delimiter
Blank Injection Month
Delimiter
Blank Injection Day
Delimiter
Blank Injection Hour
Delimiter
Blank Injection Minute
Delimiter
Method Blank Lab File ID
Delimiter
Surrogate Recovery Label
Delimiter
Surrogate Recovery Out
Delimiters
Spike Recovery Label
Delimiter
Spike Recovery Out
Delimiter
RPD Label
Delimiter
RPD Out
Delimiter
Record Sequence Number
Checksum
MM
I
Character
MINIMI!
"MB"
I
YY
I
MM
I
DD
I
HH
I
MM
I
Character
I
ft pit

I
Numeric
III
"S"
I
Numeric
I
"R"
I
Numeric
I
Numeric
Character
                                       H-28
                                                     OLM01.8  8/91

-------
      (Type 23) for Matrix Spike Duplicates (for Pesticide)

      Position:  Follows the type 20 to which it applies.
MAXIMUM
LENGTH

2
8
2
1
2
1
2
1
2
1
2
1
2
1
14
4
2
1
2
1
2
1
2
1
2
1
2
1
14
1
1
1
2
3
1
1
2
1
1
1
1
1
5
4
CONTENTS

Record Type
Delimiters
Instrument Blank Label
Delimiter
Blank Injection Year
Delimiter
Blank Injection Month
Delimiter
Blank Injection Day
Delimiter
Blank Injection Hour
Delimiter
Blank Injection Minute
Delimiter
Instrument Blank Lab File ID
Delimiters
Method Blank Label
Delimiter
Blank Injection Year
Delimiter
Blank Injection Month
Delimiter
Blank Injection Day
Delimiter
Blank Injection Hour
Delimiter
Blank Injection Minute
Delimiter
Method Blank Lab File ID
Delimiter
Surrogate Recovery Label
Delimiter
Surrogate Recovery Out
Delimiters
Spike Recovery Label
Delimiter
Spike Recovery Out
Delimiter
RPD Label
Delimiter
RPD Out
Delimiter
Record Sequence Number
Checksum
FORMAT/CONTENTS

"23"
MINIM
"IB"
I
YY
I
MM
I
DD
I
HH
I
MM
I
Character
INI
"MB"
I
YY
I
MM
I
DD
I
HH
I
MM
I
Character
I
it p it

I
Numeric
III
"S"
I
Numeric
I
"R"
I
Numeric
I
Numeric
Character
                                        H-29
                                                      OLM01.8   8/91

-------
 11.5   Format  of  the RESULTS DATA RECORD (Type 30) for Matrix Spike Duplicates


               CONTENTS                               FORMAT/CONTENTS
MAXIMUM
LENGTH
 2
 1
 1
 1
 9
 1
 9
 1
 2
 1
 3
 1
 13
 2
 1
 1
 13
 1
 1
 3
 1
 1
 5
 2
 1
 1
 1
 1
 5
 1
 1
 1
 1
 1
 5
 1
 1
 1
 13
 1
 5
4 •
               Record Type
               Delimiter
               Analyte Label
               Delimiter
               CAS Number
               Delimiter
               Internal Standard CAS Number
               Delimiter
               Concentration Unit
               Delimiter
               Result Qualifier
               Delimiter
               Results
               Delimiters
               Amount Added Label
               Delimiter
               Amount Added
               Delimiter
               CRQL Label
               Delimiter
               MS/MSD Recovery Label
               Delimiter
               MS/MSD % Recovery
               Delimiters
               RPD Label
               Delimiter
               Surrogate Recovery Label
               Delimiter
               Surrogate Recovery
               Delimiter
               Mean Concentration Label
               Delimiter
               Percent Difference Label
               Delimiter
               Percent Difference
               Delimiter
               Internal Standard Area Label
               Delimiter
               Internal Standard Area
               Delimiter
               Record Sequence Number
               Checksum
"30"
I
11C" for CAS No.
I
Numeric
I
Numeric
I
"NG"
I
STD
I
Numeric
II
A
I
Numeric
I
"U"
III
II pit

I
Numeric
II
"D"
% Recovery
I
"M"
I
II ptt

I
Numeric
I
M T It

I
Numeric
I
Numeric
Character
                                        H-30
                                                                     OLM01.8  8/91

-------
11.6  Format of the Sample Header Data Record (Type  20)  for Performance Evaluation
      Mixture
     MAXIMUM
     LENGTH
CONTENTS
FORMAT/
CONTENTS
     2
     2
     5
     1
     1
     1
     3

     2
     5
     1
     5
     I
     2
     1
     2
     1
     2
     1
     2
     1
     2
     3
     3
     3
     5
     4
"20"
Delimiters
PEM #»
Delimiter
"0"
Delimiter
"CLE"

Delimiters
CASE NUMBER
Delimiter
SDG NO.
Delimiter
YEAR OF INSTRUMENTAL ANALYSIS
Delimiter
MONTH OF ANALYSIS
Delimiter
DAY OF ANALYSIS
Delimiter
HOUR OF INSTRUMENTAL ANALYSIS
Delimiter
MINUTES OF ANALYSIS
Delimiters
ANALYTE COUNT
Delimiters
RECORD SEQUENCE NO.
CHECKSUM
Record Type
II
Sample I.D.
I
All matrices
I
Indicates Continuing Performance
Check (Pesticide Standard)
II
Numeric
I
Character
I
YY
I
MM
I
DD
I
HH
I
MM
III
Numeric
III
Numeric
Character
                                       H-31
                                                     OLM01.8   8/91

-------
11.7  Format of the Auxiliary Data Record (Type 32)  for Pesticide Calibration
      Verification Summary (Percent Breakdown Data for 4,4'-DDT from Form 7D).
     MAXIMUM
     LENGTH    CONTENTS

     2         "32"
     3         Delimiters
     2         "RT"
     1         Delimiter
     5         RT Value
     1         Delimiter
     3         "RTF"
     1         Delimiter
     5         RT Window Lower Limit
     1         Delimiter
     3         "RTT"
     1         Delimiter
     5         RT Window Upper Limit
     3         Delimiters
     2         "PB"
     1         Delimiter
     5         % Breakdown
     4         Delimiters
     5         Record Sequence No.
     4         Checksum
FORMAT/
CONTENTS

Record Type
III
Retention time label.
I
Numeric
I
First limit value
I
Numeric
I
Second limit label
I
Numeric
111
% Breakdown Label
I
Numeric (for 4,4'-DDT)
MM
Numeric
Character
                                       H-32
               OLM01.8   8/91

-------
11.8  Format of the Sample Header Data Records  (Type  20-30)  for Continuing Checks  (GC/MS
      Methods) Format

                                                      FORMAT/
                                                      CONTENTS

                                                      Record Type
                                                      II
                                                      e.g.,  VSTD050
                                                      From Exhibit B
                                                      I
                                                      All  matrices
                                                      I
                                                      Indicates
                                                      Continuing Check
                                                      I
                                                      See  Section 7.3
                                                      I
                                                      Optional
                                                      I
                                                      Character
                                                      I
                                                      YY
                                                      I
                                                      MM
                                                      I
                                                      DD
                                                      I
                                                      HH
                                                      I
                                                      MM
                                                      III
                                                      Numeric
                                                      III
                                                      Numeric
                                                      Character
                                                      FORMAT/
                                                      CONTENTS

                                                      Record Type
                                                      I
                                                      "Y"  for heated;  "N"  for non-heated
                                                      HIM
                                                      Leave  Delimiter
                                                      if none
                                                      I
                                                      Character
                                                      Illllllllll
                                                      Numeric
                                                      Character
MAXIMUM
LENGTH
2
2
12
1
1
1
3
1
3
1
5
1
5
1
2
1
2
1
2
1
2
1
2
3
3
3
5
4
MAXIMUM
LENGTH
2
1
1
5
6
1
14
11
5
4

CONTENTS
"20"
Delimiters
EPA Sample I.D.
Delimiter
"0"
Delimiter
"CLC"
Delimiter
Sample Qualifier
Delimiter
Case Number
Delimiter
SDG No.
Delimiter
Year of Instrumental Analysis
Delimiter
Month of Analysis
Delimiter
Day of Analysis
Delimiter
Hour of Instrumental Analysis
Delimiter
Minute of Analysis
Delimiters
Analyte Count
Delimiters
Record Sequence No .
Checksum

CONTENTS
"21"
Delimiter
"Y" or "N"
Delimiters
SAS NUMBER
Delimiter
LAB FILE I.D.
Delimiters
RECORD SEQUENCE NO.
CHECKSUM
                                        H-33
OLM01.8  8/91

-------
MAXIMUM
LENGTH    CONTENTS

2         "23"
1         Delimiter
1         "P"
1         Delimiter
2         INJECTION YEAR

1         Delimiter
2         INJECTION MONTH
1         Delimiter
2         INJECTION DAY
1         Delimiter
2         INJECTION HOUR
1         Delimiter
2         INJECTION MINUTES
1         Delimiter
14        DFTPP/BFB LAB FILE ID
26        Delimiters
5         RECORD SEQUENCE NO.
4         CHECKSUM
FORMAT/
CONTENTS

Record Type
I
Labels data as
"tune" data
I
Date of associated DFTPP/BFB
Injection
I
MM
I
DD
I
HH
I
MM
I
From Instrument data system
MMIMMIMM MINIMUM
Numeric
Character
MAXIMUM
LENGTH    CONTENTS

2         it 3Q.i
1         Delimiter
1         "C"
1         Delimiter
9         CAS NUMBER
1         Delimiter
9         CAS NUMBER INTERNAL STD.  UTILIZED
1         Delimiter
5         UNITS OF MEASURE
1         Delimiter
3         RESULT QUALIFIER
1         Delimiter
13        ANALYTICAL RESULT
11        Delimiters
1         "D"
7         Delimiters
13        RF PERCENT DIFFERENCE
4         Delimiters
5         RECORD SEQUENCE NO.
4         CHECKSUM
FORMAT/
CONTENTS

Record Type
I

I
Numeric
I
Numeric
I
Character
I

I
Numeric
IIIIIIIIMI
Identifies Percent Difference
MINN
Numeric
MM
Numeric
Character
                                  H-34
              OLM01.8  8/91

-------