EPA 600/4-81-055
             United States
             Environmental Protection
             Agency
&EFA     Research and
             Development
                Interim Methods For The Sampling and Analysis of
                     Priority Pollutants in Sediments
                          and Fish Tissue
             Prepared for

                Regional Guidance
             Prepared by
               Physical and Chemical Methods Branch
               Environmental Monitoring and Support Laboratory
               Cincinnati, Ohto 45268

-------
                                                               A F~T
                                                              • •—— .,"* • A*^<*
• Int.eninr.Methods for the Sampling and Analysis of
         Priority Pollutants in Sediments
                  and  Fish  Tissue
       U.  S.  Environmental  Protection  Agency
  Environmental  Monitoring  and  Support Laboratory
              Cincinnati, Ohio   45268
                    August 1977
               Revised October 1980

-------
                                   FOREWORD
    This collection of draft methods for the analyses of fish and sediment
samples for the priority pollutants was originally prepared as guidance to
the Regional Laboratories.  The intention was to update and revise the
methods as necessary if and when shortcomings and/or problems were
identified.  Some problems such as the formation of soap in the "phenol in
fish" method have been identified.  Consequently, this method has been
deleted.  Additionally, both the sediment and fish methods for volatile
organics by purge and trap analysis have been replaced.  Other editorial and
technical changes have also been made to the original methods.

    It  is the intention of the Environmental Monitoring and Support
Laboratory - Cincinnati (EMSL-Cincinnati) to improve and correct methods as
necessary.  Consequently, the user of these methods would be providing EPA a
service in calling our attention to any problems and in making suggestions
to improve the methods.  Comments should be directed to:

         Director, Environmental Monitoring and Support Laboratory
         26 W. St. Clair Street
         Cincinnati, Ohio  42568

-------
                               SAMPLE HANDLING

1.   Collection
    1.1  Samples shall  be collected according to recognized procedures.
         Preferably,  all  analyses should be performed on the same sample.  A
         minimum of 250 grams are required for the total protocol.
    1.2  The recommended  container for  the sediment sample is a standard
         one-quart, wide-mouth,  screw-cap, glass bottle with a Teflon lid
         liner.   It is  particularly important that glassware used in organic
         residue analyses be scrupulously cleaned before initial use.  At
         the time of collection,  the bottle should be filled nearly to the
         top with the sediment sample.   If the sample is collected  below a
         water column,  the threads and  sealing surfaces should be washed off
         with sample water.   "Top off"  the collected sediment sample with
         sample water and seal with the Teflon-lined screw cap.  Maximum
         effort must be made to seal the sample with a minimum of gaseous
         headspace.  The  sample must remain sealed until the aliquots for
         volatile organics are taken for analyses.
    1.3  In the case of small fish, a sufficient number should be combined
         by sampling site location and  species to obtain the minimum
         weight.  The collected samples are wrapped in aluminum foil,
         labeled with freezer tape, and placed in the freezer chest with dry
         ice.

-------
2.   Preservation
    2.1  The sediment sample should be labeled with freezer tape and
         transferred to the laboratory in an ice chest maintained at or near
         4°C.  The samples should be processed as soon as possible.
    2.2  Fish samples are to be frozen at the time of collection and must
         remain frozen until the subsamples are taken for purgeable organics.
3.   Processing
    3.1  Sediment
         3.1.1  Decant the water from the top of the sediment.  Transfer the
                sediment into a Pyrex tray and mix thoroughly with a Teflon
                spatula.  Discard sticks, stones, and other foreign objects,
                if present.  Weigh five 10.0-gram portions of the sample
                into separate 125-ml vials.  Using a crimper, tightly secure
                a septum to each bottle with an aluminum seal.  Store these
                sample aliquots in a freezer until ready for volatile
                organics analysis.
         3.1.2  Determine the percent solids in the sediment by drying a
                10-25g portion in a tared evaporating dish, overnight, at
                103°C.
                   Calculate the % solids using the equation:
                   % solids = A x 100
                              B
                   where:  A = weight of dry residue in grams
                           B = weight of wet sample in grams
         3:1.3  Transfer half of the remaining sediment sample back to the
                original sample bottle and  store  at 4°C.  This portion
                will be  used for those analyses requiring a wet sample.
                                        2

-------
            Spread the other half of the sample uniformly in the tray
            and allow to dry at room temperature for four or five days
            in a contaminant-free environment.  When dry - less than 10%
            water - grind the sample with a large mortar and pestle to a
            uniform particle size.  Discard any foreign objects found
            during grinding and transfer the powdered sediment into a
            wide-mouthed glass jar and seal with a Teflon-lined lid.
            This air-dried sample will be used for those analyses
            requiring an air-dried sample.
3.2  Fish
     3.2.1  To prepare the fish sample for analytical pretreatment,
            unwrap and weigh each fish.  Combine small fish by site and
            species until a minimum combined weight of 250g is
            obtained.  Chop the sample into 1-inch chunks using a sharp
            knife and mallet.
     3.2.2  Grind the sample using a large commercial meat grinder that
            has been precooled by grinding dry ice.  Thoroughly mix the
            ground material.  Regrind and mix material two additional
            times.  Clean out any material remaining in the grinder; add
            this to the sample and mix well.
     3.2.3  Weigh five 10.Og portions of the sample into separate 125-ml
            vials.  Using a crimper, tightly secure a septum to each
            bottle with a seal.  Store these sample aliquots in a
            freezer until ready for volatile organics analysis.
     3.2.4  Transfer the remaining fish sample to a glass container and
            store in a freezer for later subsampling and analysis.

-------
4.Special Equipment and Materials
    4.1  Ice chest.
    4.2  Wide-mouth quart bottles with Teflon lid liners.
    4.3  Teflon-coated or porcelain spatula.
    4.4  Pyrex glass tray, 8xl2x2-inch.
    4.5  Mortar and pestle (large).
    4.6  Knife, heavy blade (or meat cleaver).
    4.7  Mallet, plastic faces, 2 to 3 Ib.
    4.8  Electric meat grinder, 1/2 HP.
    4.9  Dry ice.
    4.10 Aluminum foil
    4.11 Freezer tape, for labels.
    4.12 Freezer.
    4.13 Vials,.125-ml Hypo-Vials (Pierce Chemical Co., #12995), or
         equivalent.
    4.14 Septa, Tuf-Bond  (Pierce #12720), or equivalent.
    4.15 Seals, aluminum  (Pierce #13214), or equivalent.
    4.16 Crimper,  hand (Pierce #13212), or  equivalent.

-------
               Analysis  of  Sediments  for Chlorinated  Pesticides,
               Polychlorinated Biphenyls and Non-polar Neutrals

1.  Scope
    1.1  The compounds listed in Table I are extracted from air-dried
         sediment by the Soxhlet extraction technique.  The extract is
         subsequently analyzed for pesticides and PCBs using approved
         methods (1) as cited in the Federal Register (2).  The remaining
         compounds are determined using the methods described in Appendix II
         of the Federal Register (3).  While the above referenced methods
         have been proven for pesticides and PCBs, they have not been
         sufficiently tested through extensive experimentation for the
         non-polar neutral  compounds in Table I.
2.  Special Apparatus and Materials

    2.1  Soxhlet extractor, 40-mm ID, with 500-ml round bottom flask.
    2.2  Kuderna-Danish, 500-ml, with 10-ml graduated receiver and 3-ball
         Snyder column.
    2.3  Chromatographic column - Pyrex, 20-mm ID x approximately 400-mm
         long, with coarse fritted plate on bottom.

3.  Procedure
    3.1  Extraction
         3.1.1  Weigh 30.0 grams of the previously air-dried sample into a
                tared 200-ml beaker.  Add 3 ml distilled water (10% of

-------
            sample weight),  mix well  and allow to stand for 2 hours
            while mixing occasionally.
     3.1.2  Place about 1/2" of preextracted glass wool in the bottom of
            the Soxhlet extractor chamber and quantitatively transfer
            the contents of  the beaker  into the chamber.  Place a second
            glass wool  plug  on top of the sample.  Wash the 200-ml
            beaker and  all mixing tools several times with a 1:1
            hexane/acetone mixture.   Cycle the wash mixture through the
            extractor using  a total  of 300 ml of the mixed solvent.
     3.1.3  Attach the  extractor to a 500-ml round bottom flask
            containing  a boiling stone and extract the solids for 16
            hours.
     3.1.4  After extraction is complete, dry and filter the extract by
            passing it  through a 4"  column of hexane-washed sodium
            sulfate.  Wash  the 500-ml flask and the sodium sulfate with
            liberal amounts  of hexane.   Collect the eluate in a 500-ml
            K-D evaporative  flask with a 10-ml ampul.  Concentrate the
            sample extract  to 6-10 ml.
3.2  Cleanup and Separation
     3.2.1  Adjust the sample extract volume to 10 ml and clean up the
            extraction by Florisil column chromatography according to
            the 304(g)  methodology for PCBs (1), part 10.3.  For sulfur
            removal, continue with part 10.5.3.4 of that method.  NOTE:
            If sulfur crystals are present  in the extract, separate the
            crystals from the sample by decantation.

-------
         3.2.2  Analyze the Florisil eluates for the pesticides and PCBs
                appearing in Table I, according to the approved methods (1).
         3.2.3  Analyze remaining compounds of Table I, Column C, using the
                methods described in Appendix II of the Federal Register (3)
    3.3  Standard quality assurance protocols should be employed, including
         blanks, duplicates and dosed samples as described in the
         "Analytical Quality Control Handbook" (4).  Dosing can be
         accomplished by injecting 1-20 yl of a standard into the
         homogenized sediment contained in the Soxhlet extractor chamber.
4.  Reporting of Data
    4.1  Report results in yg/kg on a dry weight basis using the percent
         moisture values determined earlier.  Report all quality control
         data with the analytical results for the samples.

-------
  Analysis of Fish for Chlorinated  Pesticides  and  Polychlorinated  Biphenyls

1.   Scope
    1.1  The  chlorinated pesticides and polychlorinated biphenyls (PCBs)
         listed  in Table I are extracted from fish using either method A or
         B as described below.  Method A employs a blender, whereas a
         Tissumizer or the equivalent is required for Method B.  Either
         procedure .results in an extract that can be incorporated directly
         into the approved procedures (1) for pesticides or PCBs as cited in
         the  Federal  Register (2).
2.   Special Apparatus and Materials
    2.1  Method  A Only
         2.1.1  Blender, high-speed - Waring Blender, Courdos, Omni-Mixer,
                or equivalent.  Explosion proof model recommended.   Quart
                container is suitable size for routine use.
         2.1.2  Buchner funnel - porcelain, 12-cm.
         2.1.3  Filter paper - 110 mm sharkskin circles.
         2.1.4  Flask, vacuum filtration - 500 ml.
    2.2  Method  B Only
         2.2.1  Tissumizer SDT-182EN (available from Tekmar Company, P. 0.
                Box 37207, Cincinnati, Ohio, 45222), or equivalent.
         2.2.2  Centrifuge - capable of handling 100 ml centrifuge tubes.
    2.3  Method  A & B
         2.3.1  Kuderna-Danish concentrator - 500 ml, with 10-ml graduated
                receiver and 3-ball Snyder column.

-------
         2.3.2  Chromatographic column - pyrex, 20 mm ID x approximately 400
                mm long,  with coarse fritted plate on bottom.
3.  Procedures
    3.1  Method A
         3.1.1  Weigh a 25 to 50g portion of frozen, ground fish and add to
                a high-speed blender.  Add lOOg anhydrous Na2S04 to
                combine with the water present and to disintegrate the
                sample.  Alternately, blend and mix with a spatula until the
                sample and sodium sulfate are well mixed.  Scrape down the
                sides of the blender jar and break up the caked material
                with the spatula.  Add 150 ml of hexane and blend at high
                speed for 2 min.
         3.1.2  Decant the hexane supernatant through a 12-cm Buchner filter
                with two sharkskin papers into a 500-ml suction flask.
                Scrape down the sides of the blender jar and break up the
                caked material with the spatula.  Reextract the residue in .
                the blender jar with two 100 ml portions of hexane, blending
                3 min. each time.  (After one min. of blending, stop the
                blender,  scrape the material from the sides of the blender
                jar, and break up the caked material between extractions.)
         3.1.3  Decant the hexane supernatants through the Buchner and
                combine with the first extract.  After the last blending,
                transfer the residue from the blender jar to the Buchner,
                rinsing the blender jar and material in the Buchner with
                three 25 to 50 ml portions of hexane.  Immediately after the
                last rinse, press the residue in the Buchner with the bottom
                of a clean beaker to force out the remaining hexane.
                                      9

-------
     3.1.4  Pour the combined extracts and rinses through a column of
            anhydrous Na2S04, 20 mm x 100 mm, and collect the eluate
            in a 500 ml Kuderna-Danish concentrator.  Wash the flask and
            then the column with small portions of .hexane and
            concentrate the extract below 10 ml.
3.2  Method B
     3.2.1  Weigh 20.Og of frozen, ground fish and add to a 100-ml
            centrifuge tube.  Add 20 ml of hexane and insert the
            Tissumizer into the sample.  Turn on the Tissumizer and
            disperse the fish in the solvent for 1 min.  Centrifuge and
            decant the solvent through a column of anhydrous Na^SO,,
            20 mm x 100 mm, and collect the eluate in a 500-ml Kuderna-
            Danish concentrator.
     3.2.2  Repeat the dispersion twice using a 20-ml aliquot each time,
            combining all dried portions of solvent in the
            concentrator.  Rinse the tissumizer and the column with
            small portions of hexane and concentrate the extract below
            10 ml.
3.3  Cleanup and Analysis
     3.3.1  Unless prior experience would indicate the fish species fat
            content is low (less than 3g per extract), the
            hexane/acetonitrile cleanup procedures described in the
            reference methods should be followed.  In all cases,
            Florisil column chromatography should be used to clean up
            the extracts before gas chromatography (1).  An electron
            capture detector is used for final measurement, and results
                                   10

-------
                are calculated in jjg/kg.  Identifications can be confirmed
                by GC/MS techniques as described in Appendix II of the
                Federal Register (3).
    3.4  Quality Control
         3.4.1  Standard quality assurance protocols should be employed,
                including blanks, duplicates, and dosed samples as described
                in the "Analytical Quality Control Handbook (4).
         3.4.2  Dose fish sample aliquots by injecting minimum amounts
                (< 20 ul total) of concentrated pest-icide or PCB solutions
                into the solid subsample 10 to 15 minutes before extraction.
4.  Reporting of Data
    4.1  Report results in ug/kg on a wet tissue basis.  Report all quality
         control data with the analytical results for the samples.
                                      11

-------
                  Analysis of Sediment for General Organics
                     by Mechanical Dispersion Extraction

1.   Scope
    1.1  This method is designed to determine solvent extractable organic
         compounds amenable to gas chromatography.   Tables I,  II, and III
         are a summary of compounds that should be extracted at an 80-100%
         efficiency.  It is a GC/MS method intended for qualitative and
         semi-quantitative determination of these compounds.  Although this
         approach has not been sufficiently tested through extensive
         experimentation, it is based on laboratory experience and is
         presently considered to be a reasonable analytical approach for
         these organic materials in sediment.
    1.2  This method is not applicable to those very volatile pollutants
         listed in Table IV.
2.   Special Apparatus and Materials
    2.1  Mechanical  dispersion device - Tissumizer
    2.2  Centrifuge - capable of handling 100-ml centrifuge tubes.
    2.3  Separatory funnels - 2 liter with Teflon stopcock.
    2.4  Sieve, 20 mesh.
3.   Procedure
    3.1  Weigh 20.Og of wet,  well-mixed sediment into a 100 ml centrifuge
         tube.  If the sediment contains grit larger than 20 mesh, it is
         necessary to extrude the sample through a 20-mesh sieve in order to
         prevent damage to the mechanical dispersion device.  Add 20 ml
                                        12

-------
     acetonitrile and insert the dispersion device into the sample.
     Disperse the sediment into the solvent for 1 min.  Centrifuge and
     decant the solvent into a 2-liter separatory funnel containing 1300
     ml of a 2 percent aqueous solution of sodium sulfate previously
     adjusted to pH 11 with 6j^ NaOH.  Repeat the dispersion twice, using
     a 20-ml aliquot each time, and combine the acetonitrile washings in
     the separatory funnel.  CAUTION:  The dispersion should be carried
     out in a fume hood to avoid exposure to acetonitrile.
3.2  Extract the aqueous acetonitrile solution in the separatory funnel
     with 60 ml of hexane for 2 min.  Drain the aqueous layer into a
     2-liter Erlenmeyer flask and pour the hexane extract through a
     short column of anhydrous sodium sulfate prerinsed with hexane.
     Collect the dried extract in a 500-ml Kuderna-Danish (K-D) flask
     fitted with a 10 ml ampul.  Repeat the extraction and drying steps
     twice, combining the extracts.
3.3  Evaporate the extract to 5 to. 10 ml in a 500-ml K-D apparatus
     fitted with a 3-ball Snyder column and a 10-ml calibrated receiver
     tube.  Allow the K-D to cool to room temperature.  Remove the
     receiver and adjust the volume to 10 ml.  Label this as the base
     neutral fraction.  If additional sensitivity is required, add fresh
     boiling chips, attach a two-ball micro-Snyder column, and carefully
     evaporate to 1.0 ml or when active distillation ceases.
3.4  Return the aqueous acetonitrile solution to the separatory funnel
     and adjust the pH with 6h[ HC1 to pH 2 or less.  Extract three times
     with 60 ml of hexane each time.  Combine the extracts, dry, and
     concentrate as described above.  Label this as the acid fraction.
                                    13

-------
    3.5  Analyze both extracts according to the methods described in
         Appendix II of the Federal Register (3).  Should the aceonitrile
         partition used in this procedure not sufficiently remove
         interferences,,florisil (1), alumina (5), and.silica gel (1),
         column chromatographic cleanup and separation techniques can be
         employed.  Sulfur can be removed by treatment with mercury (1).
    3.6  Standard quality control assurance protocols should be employed,
         including blanks, duplicates and dosed samples, as described in the
         "Analytical Quality Control Handbook" (4).  Dosing can be
         accomplished by injecting 1-20 yl of a standard solution into the
         homogenized sediment contained in the centrifuge tube.
4.  Reporting of Data
    4.1  Report results in pg/kg on a dry weight basis using the percent
         moisture values determined earlier.  Report all quality control
         data with, the analytical results for the samples.
                                      14

-------
          Analysis of  Fish  for  General  Organics  by  Solvent  Extraction
1.  Scope
    1.1  This method is designed to determine solvent extractable organic
         compounds amenable to gas chromatography.   These compounds are
         listed in Tables II and III.  It is a GC/MS method intended for
         qualitative and semi-quantitative determination of these
         compounds.   Although this approach has not been sufficiently tested
         through extensive experimentation, it is based on laboratory
         experience  and is presently the best analytical approach for these
         organic materials in fish.
2.  Special Apparatus and Materials
    2.1  Tissumizer  SDT-182EN (available from Tekmar Company, P. 0. Box
         37202, Cincinnati, Ohio 45222), or equivalent.
    2.2  Centrifuge  - capable of handling 100-ml centrifuge tubes.
    2.3  Separatory  funnels - 2-liter with Teflon stopcock. .
    2.4  Organic-free water - prepared by passing distilled water through an
         activated carbon column.
3.  Procedure
    3.1  Weigh 20.Og of ground, homogeneous fish and add to a 100-ml
         centrifuge  tube.  Add 20 ml of acetonitrile and insert the
         Tissumizer  into the sample.  Turn on the Tissumizer and disperse
         the fish into the solvent for 1 min.  Centrifuge and decant the
         solvent into a 2-liter separatory funnel containing 1300 ml of a 2
         percent aqueous solution of sodium sulfate.  Repeat the dispersion
         twice, using a 20-ml aliquot each time, and combine the
         acetonitrile in the separatory funnel.
                                      15

-------
     CAUTION:   The dispersion should be carried out in a fume hood to
     avoid exposure to acetonitrile.
3.2  Adjust the pH of the sodium sulfate/acetonitrile solution with 6N^
     NaOH to pH 11 or greater.  Use multirange pH paper for the
     measurement.  Extract the aqueous acetonitrile solution with 60 ml
     of hexane.  Shake the separatory funnel for 2 min.  Drain the
     aqueous layer into a 2-liter Erlenmeyer flask and pour the hexane
     extract through a short column of prerinsed anhydrous sodium
     sulfate.   Collect the dried extract in a 500-ml Kuderna-Danish
     (K-D) flask fitted with a 10-ml ampul.  Repeat the extraction and
     drying steps twice, combining the extracts.  Evaporate the extract
     to 5 to 10 ml in a 500-ml K-D apparatus fitted with a 2-ball Snyder
     column and a 10 ml.  Analyze by GC/MS.  If additional sensitivity
     is required, add fresh boiling chips, attach a two-ball
     micro-Snyder column, and carefully evaporate to 1.0 ml or when
     active distillation ceases.
3.3  Return the aqueous acetonitrile solution to the separatory funnel
     and adjust the pH with 6f^ HC1 to pH 2 or less.  Extract three times
     with 60 ml of hexane each time.  Combine the extracts, dry, and
     concentrate as described above.  Analyze by GC/MS.
     NOTE:  Should the partition used in this procedure not sufficiently
     remove the lipid material, gel permeation can be employed.
     (However, special expensive equipment  is necessary for this
     procedure (6)).
3.4  Standard quality assurance prptocols  should be employed, including.
     blanks, duplicates, and dosed samples, as described in the

                                    16

-------
         "Analytical  Quality Control  Handbook" (4).   Dosing can be
         accomplished by injecting 1  to 20 ul  of a standard solutuon into
         the homogenized tissue contained in a centrifuge tube.
4.  Reporting of Data
    4.1  Report results in yg/kg on a wet tissue basis.  Report all quality
         control data with the analytical results for the samples.
                                     17

-------
               Determination of Purgeable Organics  in  Sediment

1.   Scope
    1.1  This procedure is intended for use in the analysis of volatile
         organic compounds found in sediment samples (Table IV).  The
         procedure applies a modified purge/trap technique in the direct
         analysis of an undiluted sediment sample.  The method relies on the
         use of a mass spectrometer detection system, although other
         selective detectors may be used for specific compound types.
    1.2  Under ideal conditions, the minimum detectable limit has been
         determined to be 0.5 ppb.  Actual detection limits will vary due to
         sediment surfaces, water content and nonvolatile organic loading
         factors which will influence the partition coefficients of the
         volatile organics.
2.  Special Apparatus and Materials
    2.1  Tekmar LSC-1 or equivalent purge/trap apparatus.
    2.2  Septum - Teflon-faced silicone (Pierce 12722).
    2.3  Vial with  scalable cap - 20 ml (Pierce-Hypovial).
    2.4  Heating tape with temperature control.
    2.5  Hand crimper
    2.6  Standard solutions of compounds of  interest - two concentration
         levels  (10 ug/1  and 100 ug/1).
                                        18

-------
3.  Procedure
    3.1  Allow samples to equilibrate to room temperature for weighing and
         analysis.  Samples may be shipped and stored at wet ice
         temperatures; freezing is not necessary.
    3.2  Drill two holes into the septum to allow for the snug insertion of
         two 1/8" glass tubes to be used as purge gas inlet and outlet.
    3.3  The purge gas inlet should be extended to the bottom of the septum
         vial.  The purge gas outlet should extend 1/2" below the septum.
    3.4  Wrap the vial in heating tape and connect the glass tubes to the
         appropriate gas lines.
    3.5  Heat the sample at 80°C for 5 minutes.
    3.6  Withdraw the source of heat and purge the sample chamber with
         helium gas for 4 minutes at a rate of 60 ml/min.
    3.7  Desorb the trapped organics from the trap tube onto the
         chromatographic column for analysis.
    3.8  Standard quality assurance protocols should be employed, including
         blanks, duplicates and dosed, samples as described in the
         "Analytical Quality Control Handbook" (4).
4.  Calibration
    4.1  For purposes of this procedure, five sample vials of each sediment
         type must be available.
    4.2  Dose one sample vial through the septum with 10 ul of one standard
         solution.  Dose a second vial with 10 ul of the other standard
         solution.
    4.3  Proceed with the analysis of the two dosed samples and one
         non-dosed sample, using the procedure described in Section 3.
         Store the two remaining vials at 4°C for possible future analysis.
                                       19

-------
    4.4  After each  sample has been  analyzed,  proceed with data analysis.
         Subtract  the  peak areas  of  compounds  found in the undosed sample
         from the  corresponding compounds contained in the dosed sample.
         Construct a calibration  curve from the corrected dosed data;
         quantify  the  unknown.
         NOTE:  If the calculated sample concentration is greater than the
         concentration of the dosed  standard used in the dosing step, it
         will be necessary to prepare additional standards to bracket the
         unknown.
    4.5  Dry weight of the sediment  is obtained after the analysis, by first
         removing  the Teflon seal and drying the sample vial at 103-105°C
         overnight.   The gross dry weight is obtained after reequilibrat ing
         the sample vial to room temperature.   The tare weight of the vial
         is then determined after removal of the dried sediment.
5.  Reporting of Data
    5.1  Report all  results in ug/kg on a dry weight basis.  Report all
         quality control data with the analytical results for the sample.
                                      20

-------
       Analysis of Fish for Volatile Organics by Purge  and Trap  Analyses

1.  Scope
    1.1  This method is designed to determine volatile organic compounds
         amenable to purge and trap analyses.  These compounds are listed in
         Table IV.   It is a GC/MS method intended for qualitative and
         semi-quantitative determination of these compounds.
    1.2  Although the above approach has not been sufficiently tested
         through extensive experimentation, it is based, on laboratory
         experience and is presently considered to be the best analytical
         approach for volatile organic materials in fish.
2.  Special Apparatus and Materials
    2.1  Sonifier Cell Disrupter W-350 with microprobe (manufactured by
         Brawson Sonic Power Co., Danbury, Connecticut), or equivalent.
    2.2  Gas-tight syringe - 5 cc.
    2.3  Organic-free water - Prepared by passing distilled water through an
         activated carbon column.
    2.4  Standard solutions - Prepare three standard methanol solutions of
         the compounds listed in Table IV at the 50 ng/yl, 150 ng/ul, and
         300 ng/yl  concentrations.  The standard solutions should be stored
         at less than 0°C.  Solutions should be allowed to warm to room
         temperature before dosing.  Fresh standards should be prepared
         weekly.
         NOTE:  Specific GC detectors can be substituted for the MS.
    2.5  Tekmar LSC-1 or equivalent purge/trap apparatus.
    2.6  Septum - Teflon-faced silicone (Pierce 12722).
                                       21

-------
    2.7  Vial with sealable cap - 20 ml (Pierce-Hypovial).
    2.8  Heating tape with temperature control unit.
3.   Procedure
    3.1  Remove four of the sample vials containing 10.Og of homogenized
         fish from the freezer.  Open the vials and add 10 ml of organic-free
         water to each while the fish is still frozen.  Sonify the fish for
         30 sec. at maximum probe power.  Immediately reseal the vials.
    3.2  Dose one sample vial through the septum below the water level with
         10 ul of the 50 ng/ul standard methanol solution.  Dose a second
         vial with 10 ul of the 150 ng/ul standard and a third vial with 10
         ul of the 300 ng/yl standard.
    3.3  Allow samples to equilibrate to room temperature for weighing and
         analysis.
    3.4  Drill two holes into the septum to allow for the snug insertion of
         two 1/8" glass tubes to be used as purge gas inlet and outlet.  The
         purge gas inlet should be extended to the bottom of the system
         vial.  The purge gas outlet should extend 1/2" below the septum cap.
    3.5  Wrap each vial with heating tape and connect the glass tubes to the
         appropriate gas line.
    3.6  Heat each sample at 80°C for 5 minutes.
    3.7  Withdraw the source of heat and purge the sample chamber with
         helium gas for 4 minutes at a rate of 60 ml/min.
    3.8  After purging  is complete, desorb the trap tube into the gas
         chromatograph  for  analysis.  The trap column should be rapidly
         heated to 180°c and backflush with an inert gas at 20 to 60
         ml/min. for 4  minutes during this procedure.
                                      22

-------
    3.9  Analyze the undosed sample first,  followed by the 50 ng/yl dosed
         sample.  If no compounds of interest are found in the undosed
         sample and the dosed sample produces peaks to indicate recovery of
         the protocol compounds,  do not analyze the remaining samples.
         Calculate lower limits of detection based on the response obtained
         from the dosed sample.  If compounds are observed in the undosed
         sample, analyze the two  remaining  dosed samples in exactly the same
         manner.  Subtract the peak areas of compounds found in the undosed
         sample from the corresponding compounds contained in the dosed
         data; quantify the unknown.
         NOTE:  If the calculated sample concentration is greater than the
         concentration of the dosed standard used in the dosing step, it is
         necessary to prepare additional standards in order to bracket the
         unknown.  Utilize the remaining sample in the freezer for this
         purpose.
    3.10 Standard quality assurance protocols should be employed, including
         blanks, duplicates, and  dosed samples, as described in the
         "Analytical Quality Control Handbook" (4).
4.  Reporting of Data
    4.1  Report all results in yg/kg on a wet tissue basis.  Report all
         quality control data with the analytical results for the samples.
                                     23

-------
                       Analysis of Sediment for Cyanide

1.   Scope and Application
    1.1  This method is used for the determination of cyanide in sediments.
         Insoluble cyanide complexes are dissolved in 10% sodium hydroxide.
         The cyanide, as hydrocyanic acid (HCN), is released from the sample
         by means of a reflux-distillation and absorbed in a scrubber
         containing sodium hydroxide solution.  The cyanide in the absorbing
         solution is then determined colorimetrically or potentiometrically •
         or by titration.
    1.2  For cyanide levels exceeding 0.2 mg per 200 ml of absorbing liquid,
         the silver nitrate titrimetric method is used.  For cyanide levels
         below this value, the co'lorimetric procedure is used.  The probe
         method may be used for concentrations of 0.001 to 200 mg per 200 ml
         absorbing liquid.
2.   Sample Preparation
    2.1  Although a dry sample is preferred, a wet sample may also be taken
         for analysis.  In either case, the sediment samples must be well
         mixed to ensure a representative aliquot.
3.   Interferences
    3.1  Interferences are eliminated or lessened by using the distillation
         procedure.
    3.2  Fatty acids will distill and form soap under the alkaline titration
         conditions.  Therefore, acidification and extraction with
         isooctane, hexane, or chloroform is recommended.
                                      24

-------
    3.3  Ammonia and thiosulfate interfere with the electrode method
         yielding higher measurements of cyanide ion activity than are
         actually present.
4.  Preparation of Calibration Curve
    4.1  The calibration curve is prepared as described in step 8.7, (Method
         335.2), Reference 7.
    4.2  The standards must contain the same concentration of NaOH (7.1) as
         the sample.
    4.3  At least one standard should be treated as outlined below.
    4.4  The calibration curve is prepared by plotting the absorbance or the
         mv reading versus the cyanide concentration.
5.  Sample Procedure
    5.1  Place a weighed portion of the well-mixed sediment (1 to lOg) in an
         800 ml beaker with 500 ml of 10% NaOH solution and stir for 1 hour.
    5.2  Transfer the mixture to a 1-liter boiling flask.  Rinse the beaker
         with several portions of deionized distilled water and add to the
         boiling flask.
    5.3  Add 50 ml of 5% NaOH solution to the absorbing tube and dilute  if
         necessary with deionized, distilled water to obtain an adequate
         depth of liquid in the absorber.  Connect the boiling flask,
         condenser, absorber, and trap in the distillation train as shown in
         Figure 1 (Method 335.2), Reference 7.
    5.4  Add 50 ml of cone. H^SO. slowly through the air inlet tube.
         Rinse with distilled water.  Add 20 ml of Mg C12.6H20 (510 g/1)
         solution through the air inlet tube and again rinse with distilled
         water.  Continue with steps 8.4 through 8.6 (Method 335.2),
         Reference 7.
                                      25

-------
    5.5  Record the absorbance or my reading and determine the cyanide
         concentration from the calibration curve.
6.  Quality Assurance
    6.1  Initially demonstrate quantitative recovery with each distillation
         digestion apparatus by comparing distilled aqueous standards to
         non-distilled aqueous standards.  Each day, distill at least one
         standard to confirm distillation efficiency and purity of reagents.
    6.2  At least 15% of the cyanide analyses should consist of duplicate
         and spiked samples.  Quality control limits should be established
         and confirmed as described in Chapter 6 of the "Analytical Quality
         Control Handbook," Reference 4.
7.  Reporting of Data
    7.1  Report cyanide concentrations on a dry weight basis as follows:
         less than 1.0 mg/kg, to the nearest 0.01 mg/kg; 1.0 mg/kg and
         above, to two significant figures.
    7.2  Report all quality control data with the analytical results for the
         samples.
                                       26

-------
                         Analysis of Fish for Cyanide

1.  Scope and Application
    1.1  This method is used for the determination of cyanide in fish.  All
         samples must be distilled prior to the analytical determination.
         For cyanide levels exceeding 0.2 mg/200 ml of absorbing liquid, the
         silver nitrate titrimetric method is used.  For cyanide levels
         below this value, the colorimetric procedure is used.
2.  Sample Preparation
    2.1  A 5g portion of the frozen, ground fish (see "Sample Handling") is
         used for the analysis.  The sample should be thawed before the
         analysis begins.
3.  Preparation of Calibration Curve
    3.1  The calibration curve is prepared from values for portions of
         spiked fish tissue distilled in the manner used for the tissue
         sample being analyzed.  For preparation of the calibration
                                                   i
         standards, choose and weigh a 50g portion of fish and blend in a
         Waring blender (or equivalent) with 10 ml of 10% NaOH and
         sufficient deionized, distilled water to bring the volume of the
         mixture to 500 ml.
    3.2  Using a volumetric pipet which has had the tip removed, withdraw
         eight 50 ml portions and place in a series of 1 liter boiling
         flasks.  Seven of the flasks should be spiked with progressively
         larger volumes of the cyanide standard as given in 8.7 (Method
         335.2), Reference 7.  Adjust the final volume of each flask to 500
         ml with deionized, distilled water.
                                      27

-------
    3:3  Add 50 ml  of 5% NaOH solution to the absorbing tube and dilute, if
         necessary,  with deionized distilled water to obtain an adequate
         depth of liquid in the absorber.  Connect the boiling flask,
         condenser,  absorber, and trap in the train as.shown in Figure 1
         (Method 335.2), Reference 7.
    3.4  Continue with step 8.2 through 8.6 (Method 335.2), Reference 7.
    3.5  The calibration curve is prepared by plotting the absorbance versus
         the cyanide concentration.  The blank absorbance value must be
         subtracted from each value before plotting the curve.
4.  Sample Procedure
    4.1  Place a weighed portion of the ground fish (approximately 5g) in a
         blender with 100 ml of deionzized, distilled water and 1 ml of 5%
         NaOH solution.
    4.2  Blend until a homogeneous mixture is obtained and transfer to a
         1-liter boiling flask.  Rinse the blender with several portions of
         deionized, distilled water totaling 400 ml and add to the boiling
         flask.
    4.3  Add 50 ml of 5% NaOH solutuion to the absorbing tube and dilute if
         necessary with deionized, distilled water to obtain an adequate
         depth of liquid in  the absorber.  Connect the boiling flask,
         condenser, absorber, and trap in the distillation train as shown in
         Figure  1 (Method 335.2)  and continue with step 8.2 through 8.6,
         Reference 7.
    4.4  Read the absorbance and  determine the cyanide concentratjion from
         the  calibration curve.
                                       28

-------
5.  Quality Assurance
    5.1  Initially, demonstrate quantitative recovery with each distillation
         digestion apparatus by comparing distilled aqueous standards to
         non-distilled aqueous standards.  Each day, distill at least one
         standard to confirm distillation efficiency and purity of reagents.
    5.2  At least 15% of the cyanide analyses should consist of duplicate
         and spiked samples.  Quality control limits should be established
         and confirmed as described in Chapter 6 of the "Analytical Quality
         Control Handbook," Reference 4.
6.  Reporting of Data
    6.1  Report cyanide concentrations as follows:  less than  1.0 mg/kg, to
         the nearest 0.01 mg; 1.0 mg/kg'and above, to two significant
         figures.
    6.2  Report all quality control data with the analytical results for the
         samples.
                                      29

-------
                       Analysis of Sediment for Phenols

1.  Scope and Application
    1.1  This method is used for the determination of phenolics  in
         sediments.  All samples must be distilled prior to the
         determination of phenols, using the procedure given on  page 576,
         Reference 8.  Use Method 510B for samples that contain  less than  1
         mg phenol/kg and method 510C for samples that contain more than 1
         mg phenol/kg.
2.  Sample Preservation and Preparation
    2.1  Biological degradation is inhibited by cooling the sample to 4°C.
         If the sample cannot be analyzed within 24 hours, it should be
         frozen.
    2.2  A 5g .portion of the wet, or air-dried sediment is used  for the
         analysis.  If the sample has been frozen, it should be  thawed
         before the analysis begins.
3.  Preparation of Calibration Curve
    3.1  The calibration curve is prepared as described on p. 579, 4.a.3
         (Ref. 8) for samples containing less than 1 mg/kg and p. 581 for
         samples above  1 mg/kg.
    3.2  Record the absorbance of the standards and plot the values against
         micrograms of phenol.
4.  Sample Procedure
    4.1  Place a 5g portion of the wet, or air-dried sediment into a 200 ml
         beaker with  100 ml of distilled water.  Mix well and lower the pH
         to 4.0 with  (1 + 0) H3P04 using a pH meter.
                                     30

-------
    4.2  Add 5 ml of 10% CuSCL solution, mix and transfer to a 1-liter
         distilling flask.
    4.3  Rinse the beaker with several portions of distilled water and add
         to the distilling flask.  Adjust the volume in the flask to 500 ml.
    4.4  Using a 500-ml graduated cylinder as a receiver, begin the
         distillation as described on p. 577, Method 510A; 4b, Reference 8.
    4.5  Continue with the procedure using either the Chloroform Extraction
         Method 510B, p. 577, Reference 8, or the Direct Photometric Method
         510C, p. 580, Reference 8.
    4.6  Record the absorbance and determine the micrograms of phenol from
         the appropriate calibration curve.
5.  Quality Assurance
    5.1  Demonstrate quantitative recovery with each distillation apparatus
         by comparing aqueous distilled standards to non-distilled
         standards*  Each day, distill at least one standard to confirm the
         distillation efficiency and purity of reagents.
    5.2  At least 15% of the phenol analyses should consist of duplicate and
         spiked samples.  Quality control limits should be established and
         confirmed as described  in Reference 4.
6.  Reporting of Data
    6.1  Report phenol concentrations on a dry weight basis as follows:
         Method 510B, to the nearest yg/kg
         Method 510C, for less than 1.0 yg/kg to the nearest 0.01 yg and for
         1.0 mg/kg and above to  two significant figures.
    6.2  Report all quality control data when reporting results of sample
         analysis.
                                      31

-------
                      Analysis of Sediment for Mercury

1.  Scope  and  Application
    1.1  This  method  is used for the determination of total  mercury (organic
        and  inorganic) in sediment.  A weighed portion of the sample is
        digested with aqua regia for 2 minutes at 95°C followed by
        oxidation with potassium permanganate.  Mercury is  subsequently
        measured by  the cold vapor technique.
    1.2  The  range of the method is 0.2 to 5 yg/g but may be extended above
        or below the normal range by increasing or decreasing sample size
        or through instrument and recorder control.
    1.3  For  a complete description of the method, the reader is referred to
        "Methods for Chemical Analysis of Water and Waste"  (7), Method
        245.5.
2.  Sample Preparation
    2.1  Although a wet sample may be taken for analysis, a  dry sample
        provides for ease of handling, better  homogeniety,  and better
        storage.
3.  Preparation  of Calibration Curve
    3.1  The  calibration curve is prepared using distilled water standards,
        treated in the same manner as the sediment samples  being  analyzed.
        Plot  peak height versus the mercury concentration.   The peak height
        of the  blank is subtracted from each of the other values.
4.  Sample Procedure
    4.1  Weigh 0.2 to 0.3g portions of the dry sample and place in  the
        bottom of a  BOD bottle.  (If a wet sample is to be  analyzed, a
                                     32

-------
         proportionately larger sample must be taken.)  Add 5 ml of
         distilled water and 5 ml  .of aqua regia and place the bottle in a
         water bath maintained at 95°C for 2 minutes.
    4.2  Cool, add 50 ml distilled water, 15 ml of potassium permanganate
         solution and return the bottle to the water bath for an additional
         30 minutes.  Add additional KMnO., as necessary, to maintain
         oxidizing conditions.
    4.3  Continue with the procedure as described.
5.  Calibration
    5.1  Measure the peak height of the unknown from the chart and read the
         mercury value from the standrd curve.
    5.2  Calculate the mercury concentration in the sample by the formula:
    5.3  Report mercury concentrations on a dry weight basis as follows:
         Below 0.1 yg/gm, < 0.1 yg; between 0.1 and 10 yg/gm, to nearest
         0.01 yg; above 10 yg/gm, to nearest yg.
6.  Quality Assurance
    6.1  Standard quality assurance protocols should be employed, including
         blanks, duplicates, and spiked samples, as described in the
         "Analytical Quality Control Handbook"  (4).
7.  Precision and Accuracy
    7.1  The following standard deviations on replicate sediment samples
         were recorded by a single operator at  the indicated levels:  0.29
         yg/gm±0.02 and 0.82 yg/gm±0.03.  Recovery of mercury at these
         levels, added as methyl mercuric chloride, was 97% and 94%,
         respectively.
                                      33

-------
                        . Analysis of Fish for Mercury

1.  Scope and Application
    1.1   This method is used for determination of total mercury (organic and
         inorganic) in fish. A weighed portion of the sample is digested
         with sulfuric and nitric acid at 58°C followed by overnight
         oxidation with potassium permanganate at room temperature.  Mercury
         is subsequently measured by the conventional cold vapor technique.
    1.2   The range of the method is 0.2 to 5 yg/g but may be extended above
         or below the normal instrument and recorder control..
2.  Sample Preparation
    2.1   The sample may be prepared as described under "Sample Handling" or
         the special metal procedure may be used.  A 0.2 to 0.3g portion
         should be taken for each analysis.  The sample should not be
         allowed to thaw before weighing.
3.  Preparation of Calibration Curve
    3.1   The calibration curve is prepared from values for portions of
         spiked fish tissue treated in the manner used for the tissue
         samples being analyzed.  For preparation of the calibration
         standards, choose a 5g portion of fish and blend in a Waring
         blender.
    3.2  Transfer accurately weighed portions to each of six dry BOD
         bottles.  Each sample should weigh about 0.2 grams.  Add 4 ml of
         cone. HUSO, and 1 ml of cone. HN03 to each bottle and place
         in water bath at 58°C until the tissue is completely dissolved
         (30 to 60 min.).
                                     34

-------
    3.3  Cool and transfer 0-, 0.5- 1.0-, 2.0-, 5.0- and 10.0- ml aliquots
         of the working mercury solution containing 0 to 1.0 ug of mercury
         to the BOD bottles.  Cool to 4°C in an ice bath and cautiously
         add 15 ml of potassium permanganate solution.  Allow to stand
         overnight at room temperature under oxidizing conditions.
    3.4  Add enough distilled water to bring the total volume to
         approximately 125 ml.  Add 6 ml of sodium chloride-hydroxylamine
         sulfate solution to reduce the excess permangante.
    3.5  Wait at least 30 sec. after the addition of hydroxylamine.
         Treating each bottle individually, add 5 ml of.the stannous sulfate
         solution and immediately attach the bottle to the aeration
         apparatus.
    3.6  Continue with the procedure as given  in Method 245.1 for water  (7).
         The calibration curve is prepared by  plotting the peak height
         versus the mercury concentration.  The peak height of the blank is
         subtracted from each of the other values.
4.  Sample Procedure
    4.1  Weigh 0.2 to 0.3g portions of the sample and place in the bottom of
         a dry BOD bottle.  Care must be taken that none of the sample
         adheres to the side of the bottle.  Add 4 ml of cone. H2S04 and
         1 ml of cone. HNO, to each bottle and place in a water bath
         maintained at 58°C until the tissue is completely dissolved (30
         to 60 minutes).
    4.2  Cool to 4°C in an ice bath and cautiously add 5 ml of potassium
         permanganate solution in 1 ml increments.  Add an additional 10 ml
         of more of permangante, as necessary  to maintain oxidizing

                                      35

-------
         conditions.  Allow to stand overnight at room temperature (see
         NOTE).   Continue as described under 3.4.
         NOTE:   As an alternate to the overnight digestion, the
         solubilization of the tissue may be carried out in a water bath at
         80°C for 30 min.  The sample is then cooled and 15 ml of
         potassium permanganate solution added cautiously.   At this point,
         the sample is returned to the water bath and digested for an
         additional 90 min. at 30°C (9).  If this method is followed, the
         calibration standards must also be treated in this manner.
         Continue as described under 3.4.
5.  Calculation
    5.1  Measure the peak height of the unknown from the chart and read the
         mercury value from the standard curve.
    5.2  Calculate the mercury concentration in the sample by the formula:
    5.3  Report mercury concentrations as follows:
         Below 0.1 ug/gm, < 0.1 ug; between 0.1 and 1 ug/gm, to nearest 0.01
         yg; between 1 and 10 ug/gm, to nearest 0.1 ug; above 10 ug/gm, to
         nearest ug.
6.  Quality Assurance
    6.1  Standard quality assurance protocols should be employed, including
         blanks, duplicates, and spiked samples as described in the
         "Analytical Quality Control Handbook" (4).
    6.2  Report all quality control data when reporting resaults of sample
         analyses.
                                      36

-------
7.  Precision and Accuracy
    7.1  The following standard deviations on replicate fish samples were
         recorded at the indicated levels:  0.19 yg/gm±0.02, 0.74
         ug/gm±0.05, and 2.1 yg/gm±0.06.  The coefficients of variation at
         these levels were 11.9%, 7.0%, and 3.6%, respectively.  Recovery of
         mercury at these levels, added as methyl mercuric chloride, was
         112%, 93%, and 86%, respectively.

-------
                       Analysis of Sediments for Metals

1.   Scope and Application
    1.1  This method is used for the determination of antimony, beryllium,
         cadmium, chromium, copper, lead, nickel, silver, thallium, and zinc
         in sediments.
2.   Summary of Method
    2.1  The sediment is prepared for analysis by drying and grinding the
         sample.  A representative portion is subjected to wet oxidation-
         digestion prior to analysis by atomic absorption.
    2.2  For a discussion of basic principles, general operating parameters,
         preparation of standards and calibration, and the method of
         standard addition, the reader is referred to "Methods for Chemical
         Analysis of Water and Wastes" (7) and the individual methods as
         follow:
                                      38

-------
Page References to "Methods for Chemical Analysis of Water and Wastes, 1979-   '
ELEMENT   Ag     Be     Cd     Cr     Cu     Ni     Pb     Sb     Tl    Zn  -
METHOD    27271  270TI211712T£7I22"0~7I27571  21STTIZoTH279.1 289.1
3.   Preservation and Handling
    3.1  The sample should be stored at 4°C if the analysis can be carried
         out within 7 days of collection.  For longer periods, the samples
         should be frozen.  An alternative is to dry the sample as soon as
         possible, grind it with a mortar and pestle removing rocks, sticks,
         and other foreign objects and store the sediment in a vial or other
         suitable container.
    3.2  Dust in the laboratory environment, impurities in reagents, and
         impurities on laboratory apparatus, which the sample contacts, are
         all sources of potential contamination.  All glassware should be
         thoroughly washed with detergent and tap water, rinsed with 1:1
         nitric acid, tap water, and finally deionized, distilled water in
         that order.  NOTE:  Chromic acid may be useful to remove organic
         deposits from glassware; however, the analyst should be cautioned
         that the glassware must be thoroughly rinsed with water to remove
         the last trace of chromium.  This is especially important if
         chromium is to be included in the analytical scheme.  A commercial
         product - NOCHROMIX - available from Godax Laboratories, 6 Varick,
         New York, NY 10013, can be used in place of chromic acid.
4.  Sample Preparation
    4.1  Dry a representative portion of the well-mixed sample (10 to 25g)
         at 60°C until all moisture has been removed.
                                       39

-------
    4.2  Grind the dry sample with a mortar and pestle, removing sticks,
         stones, and other foreign material.  Store the sample in glass or
         plastic vials removing aliquots as needed.
5.  Procedure
    5.1  Weigh l.OOg of the well-mixed sediment into a 250 ml Erlenmeyer
         flask and add 50 ml deionized water, 0.5 ml HN03 (sp. gr. 1.42)
         and 5 ml of HC1 (sp. gr. 1.10) to each flask.
    5.2  Heat the samples, blanks and standards on a hotplate maintained at
         approximately 95°C until the volume has been reduced to 15 to 20
         ml, making certain that the samples do not boil.
    5.3  Cool and clarify the sample by centrifugation or by filtration
         through Whatman No. 42 filter paper or equivalent.
    5.4  Dilute the sample to 100 ml or some appropriate volume based on the
         concentration present.
    5.5  Proceed with the appropriate method for the atomic absorption
         analysis of the metals of interest.
6.  Calibration
    6.1  From the values read off the appropriate calibration curve,
         calculate the concentration of each metal pollutant in the sediment
         as follows:
                mg/1 of constituent         volume of prepared
                prepared sample        X         sample in ml
         mg/kg =	        	•      	•__
                            weight of dry sample in g
7.  Quality Assurance
    7.1  Standard quality assurance protocols should be employed, including
         blanks, duplicates, spiked and samples as described in the
         "Analytical Quality Control Handbook" (4).
                                     40

-------
7.2  Report all quality control data when reporting results of sample



     analyses.
                                 41

-------
                         Analysis of Fish for Metals

1.  Scope
    1.1  This method is used for the determination of antimony, arsenic,
         beryllium, cadmium, chromium,  copper,  lead, nickel, selenium,
         silver, thallium,  and zinc in  fish tissue.
2.  Summary of Method
    2.1  The fish is prepared for analysis by being chopped into small
         pieces, homogenized in a blender with  dry ice, and solubilized by
         either dissolution after dry ashing or a wet oxidation digestion.
         After sample preparation,  atomic absorption - either direct
         aspiration, gaseous hydride, or a fTameless technique - is used to
         measure the concentration  of the pollutant.
3.  Preservation and Handling
    3.1  Although an aliquot of the ground fish as prepared under "Sample
         Handling" may be used for  the  metals determination, it may be more
         desirable to prepare an individual  fish to avoid possible metal
         contamination from the grinder.  Dust  in the laboratory
         environment, impurities in reagents, and impurities on laboratory
         apparatus that the sample  contacts are all sources of potential
         contamination.  All glassware  should be thoroughly washed with
         detergent and tap water, rinsed with 1:1 nitric acid, then tap
         water, and finally deionized,  distilled water.
         NOTE:  Chronic acid may be useful to remove organic deposits from
         glasware; however, the analyst should be cautiuoned that the
         glassware must be thoroughly rinsed with water to remove the last

                                     42

-------
         trace of chromium.  This is especially important if chromium is to
         be included in the analytical scheme.  A commercial product
         - NOCHROMIX - available from Godax Laboratories, 6 Varick Street,
         New York, NY, 10013, can be used in place of chromic acid.
4.  Sample Homogenization
    4.1  If a fish sample other than that prepared under "Sample Handling"
         is to be used for metals analyses, unwrap and weigh the frozen fish
         at the time of processing.  Select a fish that weighs between 50
         and 300g.  If an analysis is required for a fish, < 300g, a 50g
         representative portion must be taken from the sample after it has
         been pretreated as described in "Sample Handling" on page 1 of this
         document.
    4.2  After weighing, the fish should be chopped into approximately 1-in.
         or smaller chunks with a meat cleaver or a knife and mallet (2 to
         3-lb).  Smaller pieces ensure efficient grinding.
    4.3  Place crushed or pelleted dry ice into the blender container.  The
         weight of dry ice should be equal to, or greater than, the weight
         of the fish.
    4.4  Turn on the  blender for  10 sec. to pulverize the ice and chill the
         blender.
    4.5  Add the pieces of fish and blend at high speed until the mixture is
         homogeneous.  This usually requires 2 to 5 minutes.  Add more dry
         ice  if needed to  keep the fish frozen.
    4.6  Pour the homogenate into a plastic bag and close the bag with a
         rubber band.  Do  not  seal the bag tightly to allow COp to escape.
                                      43

-------
    4.7  Place the bag in the freezer (-12°C for at least 16 hr.) until
         ready to proceed with the digestion step.
         NOTE:  If desired, the blender blades can be modified in order to
         have the leading edge of the blades (the sharpened edge) turned
         down so that, as it rotates, the blade will throw the material
         upwards.  Stainless steel blades may be a possible source of nickel
         and chromium contamination and should be noted if detected.  If a
         tantalum blade is available, it should be substituted for the
         stainless steel.
         The hole in the blender lid should be enlarged sufficiently to
         allow the evolved gas to escape.  Hold a cloth or labwipe over this
         hole when blending to prevent loss of the sample material.  A glove
         should be worn to prevent possible freezing of the skin by escaping
         gas.
5.  Reagents
    5.1  Deionized, distilled water:  Prepare by passing distilled water
         through a mixed bed of cation and anion exchange resins.  Use
         deionized, distilled water for the preparation of all reagents and
         calibration standards and as dilution water.
    5.2  Nitric acid (cone.):  If metal impurities are present, distill
         reagent grade nitric acid in a borosilicate glass distillation
         apparatus.
    5.3  Sulfuric acid, ACS grade (95.5% to 96.5%).
    5.4  Sulfuric acid - 20% V/V solution.  Carefully add 200 ml of
         concentrated ^SO^ to 500 ml of water.  Cool and dilute to 1
         liter with water.
                                     44

-------
    5.5  Hydrochloric acid, ACS grade.
    5.6  Hydrogen Peroxide, 50% stabilized ACS grade.
    5.7  Dry ice (frozen carbon dioxide), pellet form preferred.
6.  Apparatus
    6.1  Blender, Waring, two-speed, stainless steel blade or tantalum
         blade, if available, glass container capacity 1000 ml, or
         equivalent equipment.
    6.2  Drying oven - Controllable with the range of 100° to 150°C with
         less than ±5°C variation.  Check calibration of oven temperature
         control to ensure accurate ashing temperatures.  Furnace must be
         operated in suitable fume hood.
    6.3  Hot plate, controllable within the range of 80°C to 400°C.  Hot
         plate must be operated in fume hood.
7.  Procedure
    Except for mercury, which requires a cold vapor technique, the metals
    can be divided into two groups for continued processing.
         Group I:   Be, Cd, Cr, Cu, Pb, Ni, Ag, Tl, and Zn.
         Group II:  As and Se.
    Group I is digested by a dry ashing process (11) with the use of an
    ashing aid; Group II is prepared utilizing a wet ashing process.
    7.1  Group I - Metals
         7.1.1  Remove the homogenized sample from the freezer and weigh
                approximately lOg into a tared, 100-ml tall form, Pyrex
                beaker.  Subtract the beaker weight from the total and
                record the wet sample weight.
                                     45

-------
7.1.2  Add 25 ml of 2Q% sulfuric acid.  Mix each sample thoroughly
       with a glass stirring rod ensuring all sample material is
       wetted by the acid.   Rinse the stirring rod with water into
       the ashing vessel and cover the sample with a ribbed watch
       glass.
7.1.3  Heat the samples in an oven or furnace at 110±5°C until a
       charred viscous sulfuric acid/sample residue remains.
       Usually 12 to 16 hrs. (overnight) is sufficient.  Transfer
       the ashing vessels containing the samples to a cold, clean
       muffle furnace which is provided with good external
       ventilation (fume hood), ensuring that the sample remains
       covered during the transfer.  Initially set the furnace at
       125 C and increase the temperature approximately every
       hour  in 50° increments up to 275°C.  Hold the
       temperature at 275°C for 3 hrs.  Finally, increase the
       temperature to 450°C (at 50° per hour) and hold for  12
       to 16 hrs.  (overnight).  Remove the covered ashing vessels
       from  the furnace and allow to cool to room temperature in a
       clean, draft-free area.
7.1.4  After initial overnight ashing, some residual carbon may
       remain in the samples.  Treat each sample ash with 0.5 ml of
       water and 1 ml of concentrated nitric acid (whether  or not
       they  are already white).  Evaporate carefully just to
       dryness  on  a warm hotplate  (in a fume hood).  Place  the
       ashing vessels  (covered with watch glasses)  in  a cool muffle
       furnace  and raise the temperature  to  300°C and  hold  for
                            46

-------
       exactly 30 min.  Remove each covered sample ash from the
       furnace and allow to cool as before.  If residual carbon
       remains, repeat the nitric acid treatment until a carbon-
       free white ash is obtained.  The covered ashing vessels
       containing the ash may be stored in a dessicator or in a
       laminar flow clean hood.
       NOTE:  Copious carbon residues (i.e., black ashes) after
       overnight ashing may indicate inefficient or uneven heating
       within the furnce.  Routine calibration of the furnace is
       advised.
7.1.5  Add 0.5 ml of nitric acid and 10 ml of water to each cool
       ashing vessel, then warm gently on a hotplate at 80 to
       90°C for 5 to 10 min. to effect dissolution of the ash.  A
       small amount of insoluble white siliceous-like residue may
       remain undissolved; do not filter the residue because of the'
       possibility of contamination.  Quantitatively transfer the
       contents of each ashing vessel into a 100 ml volumetric
       flask, dilute to volume with water, and shake thoroughly.
       Allow any residue to settle to the bottom of the flask
       (about 2 hr).  Do not shake the sample further before taking
       an aliquot.  The sample is now ready for analysis.
       NOTE:  The presence of a precipitate other than the
       insoluble siliceous-like material may result in low or
       eratic results for Pb.  Precipitate formation can result
       from heating the samples too long or at too high a
       temperature after nitric acid treatment of the ash.

                            47

-------
                Precipitate formation must be avoided by maintenance of
                appropriate ashing temperatures.
         7.1.6  The prepared sample should be analyzed by atomic
                absorption.  For a discussion of basic principles, the
                method of standard addition, the chelation/solvent
                extraction procedures, general instrumental operating
                parameters, and preparation of standards and calibration see
                the section on "Atomic Absorption Methods" (7), and the
                individual analyses sheets as follow:
ELEMENT   Ag     Be     Cd     Cr     Cu     Ni     Pb     Sb     T1    Zn  -
METHOD    272.1  210.1  213.1  218.1  220.1  249.1  239.1  204.1  279.1 289.1
         7.1.7  Because of the adequate sensitivity by conventional flame AA
                and the expected concentration levels of cadmium, copper,
                and zinc in the sample, these three elements should be
                analyzed by direct aspiration.  The furnace technique is
                preferred for the analysis^of the other Group I metals
                because of their expected  low concentrations.  When using
                the furnace technique, the operating parameters and
                instructions as specified  by the particular instrument
                manufacturer should be followed.  If the concentration
                detected by the furnace procedure is beyond the working
                range of the standard curve, the sample should be either
                diluted and reanalyzed or  analyzed by direct aspiration.
                The method of standard additions should be employed when
                needed.  If the sample matrix is so complex that  sample
                dilution followed by furnace analysis cannot be used, or if
                the use of the chelation/solvent extraction techniqe for
                                      48

-------
            concentration of Ag, Ni, Pb, and Tl is preferred, the
            procedure as described in Methods for Chemical Analysis of
            Water and Wastes, Reference 7, should be utilized.
7.2  Group II - Metals
     7.2.1  Remove the homogenized sample from the freezer and weigh
           . approximately 5g into a tared, 120-ml conical beaker.
            Subtract the beaker weight from the total and record the wet
            sample weight.
     7.2.2  Add 5 ml of cone. HNO.,.  Then slowly add 6 ml of cone.
            FLSCK and cover with a watch glass.
     7.2.3  Place beaker on hot plate and warm slightly.  Continue
            heating until the mixture becomes dark or a possible
            reducing condition is evident.  Do not allow the mixture to
            char.  Remove beaker from hotplate and allow to cool.
            NOTE:  Remove beaker if foaming becomes excessive.
     7.2.4  Add an additional 5 ml of cone. HNO,, cover with a watch
            glass, and return beaker to hotplate.  Repeat step 7.2.3.
     7.2.5  When mixture again turns brown, cool, and slowly add 5 ml of
            50% hydrogen peroxide.  Cover with watch glass and heat
            gently until the initial reaction has ceased.  If the
            solution becomes dark, repeat the peroxide addition, several
            times if necessary, and heat to SO., fumes.  If charring
            occurs, add further 1 ml portions of hydrogen peroxide until
            the fuming sulfuric acid remains colorless or very light
            yellow.  (If at any stage it appears that the sulfuric acid
            may apprpach dryness, cool, add 2 to 3 ml of sulfuric acid,
            and continue.)
                                 4S

-------
         7.2.6  Cool,, add 40 ml of cone. HC1 and dilute to 100 ml with
                deionized, distilled water.  The sample is now ready for
                analysis.
         7.2.7  The Group II metals should be analyzed by atomic absorption
                using the gaseous hydride technique.  The apparatus setup,
                standard preparation and calibration, and analysis procedure
                that is to be followed is given, starting on Page 159,
                Reference 8.  From the prepared sample, a 25-ml aliquot
                should be withdrawn and the analysis continued as described
                in Section 3.d, Page 162, Reference 8.
8.  Calculation
    8.1  Using the values from the appropriate calibration curve, calculate
         the concentration of each metal pollutant in the fish as follows:
         If the concentration of standards in the calibration curve.is
         plotted as mg/1,
                       mg/1 of constituent          volume of- prepared
                       in prepared sample      X    sample in ml
         ug/gram =     	       	^__^	
                                  weight of wet sample in g
         If the concentration of standards in the calibration curve is
         plotted as yg/1,
                        yg/1 of constituent          volume of prepared
                        in prepared sample     X     sample in ml
         yg/gram =      		    ^	
                                  weight of wet sample in g

9.  Quality Assurance
    9.1  Standard quality assurance protocols should be employed, including
         blanks, duplicates, and spiked samples, as described in the
         "Analytical Quality Control Handbook (4).
                                      50

-------
9.2  Report all quality control data when reporting results  of  sample
     analyses.
                                  51

-------
                Analysis  of  Sediment  for  Arsenic  and  Selenium

1.   Scope and Application
    1.1  This method is to be used for the determination of arsenic and
         selenium in sediment.  A weighed portion of the wet, well-mixed
         sediment is digested with HNO, and HpSCL followed by
         treatment with HpOp.  Arsenic and selenium are subsequently
         determined by the gaseous hydride technique.
    1.2  The range of the method is 	 to	yg/g but may be extended
         by varying the sample size.
2.   Sample Preparation
    2.1  The analysis should be performed on a wet, well-mixed sample.
3.   Preparation of Calibration Curve
    3.1  The calibration curve is prepared using distilled water standards,
         treated in the same manner as the samples being analyzed.
4.   Procedures
    4.1  Weigh approximately 5 grams of the wet, well-mixed, sediment  into  a
         tared 125-ml conical beaker.
    4.2  Add 5 ml of cone. HNO,.  Then slowly add 6 ml cone. H,,S04 and
         cover with a watch glass.
    4.3  Place the beaker on hot plate and warm slightly.  (NOTE:  Remove
         beaker if foraming becomes excessive.)  Continue heating until the
         mixture becomes dark or a possible reducing condition is evident.
         Do not allow the mixture to char.  Remove the beaker from the
         hotplate and allow to cool.
    4.4  Add an additional 5 ml of cone.  HNO,, cover with a watch glass,
         and return beaker to hot plate.  Repeat step 4.3.

                                      52

-------
    4.5  When mixture  again turns brown, cool, and slowly add 5 ml of 50%
         hydrogen  peroxide.  Cover with watch glass and heat gently until
         the initial  reaction has ceased.  If the solution becomes dark,
         repeat the peroxide addition, several times if necessary, and heat
         to SCU fumes.   If charring occurs, add additional 1 ml portions
         of hydrogen  peroxide until the fuming sulfuric acid remains
         colorless or  very light yellow.  (If at any stage the sulfuric acid
         approaches dryness, cool, add 2 to 3 ml of additional sulfuric
         acid, and continue.)
    4.6  Cool, add 40  ml of cone. HC1 and dilute to 100 ml with deionized,
         distilled water.   The sample is now ready for analysis by the
         gaseous hydride technique.
    4.7  The apparatus setup, standard preparation and calibration, and
         analytical procedure to be followed is given beginning on page 159,
         Reference 8.   A 25-ml aliquot should be withdrawn from the prepared
         sample and the analysis continued as described in Section 3.d, page
         162, Reference 8.
5.  Calibration
    5.1  Calculate the concentration of arsenic and selenium present in
         mg/kg on a dry weight basis.
6.  Quality Assurance
    Standard quality assurance protocols should be employed, including
    blanks, duplicates, and spiked samples as described in the "Analytical
    Quality Control Handbook" (4).
                                      53

-------
                                  REFERENCES

1.  "Methods for Benzidine, Chlorinated Organic Compounds, Pentachlorophenol
    and Pesticides in Water and Wastewater," U. S. Environmental Protection
    Agency, Environmental Monitoring and Support Laboratory, Cincinnati,
    Ohio, 45268, 1978.
2.  Federal Register, Volume 41, No. 232, p. 52780, Wednesday, December 1976.
3.  Federal Register, Volume 44, No. 233, p. 69464, Monday, December 3, 1979.
4.  Handbook for Analytical Quality Control in Water and Wastewater
    Laboratories.  U. S. Environmental Protection Agency, Technology
    Transfer, 1979.
5.  Boyle, H. W.,.et al., Adv. Chem. Serv., 60, 207 (1966).
6.  Stalling, D. L.; R. C. Tindle; J. L. Johnson;. "Cleanup of Pesticide and
    Polychlorinated Biphenyl Residues in Fish Extracts by Gel Permeation
    Chromatography." JOAC, 155, 32-38 (1972).
7.  "Methods for Chemical Analysis of Water and Wastes," U. S. Environmental
    Protection Agency, Technology Transfer.  (1979).
8.  "Standard Methods for the Examination of Water and Wastewater," 14th
    edition  (1975).
9.  Bishop,  J. N., "Mercury in Fish," Ontario Water Resources Comm.,
    Toronto, Ontario, Canada, 1971.
10. Jones, J. W.; R. J. Gajan; K. W. Boyer; J. A. Fiorino; "Dry Ash -
    Voltammetric Determination of Cadmium, Copper, Lead, and Zinc in
    Foods."  JOAC, 60, 826.  (1977).
11. Jones, J. W.; R. J. Gajan, K. W. Boyer; J. A. Fiorino; "Dry Ash -
    Voltammetric Determination of Cadmium, Copper, Lead, and Zinc in Foods."
    JOAC, 60, 826.   (1977).
                                      54

-------