oEPA
          United States
          Environmental Protection
          Agency
                 Office of Watet
                 Engineering and Analysis Division (WH-552)
                 Washington, DC 20460
EPA821-R-92-008
December 1992
Methods for the Determination of
Diesel, Mineral, and Crude Oils in
Offshore Oil and Gas Industry
Discharges
                                           Printed on Recycled Paper

-------
Acknowledgements


      The methods in this compendium were prepared under the direction of William A. Telliard of the
U.S. Environmental  Protection Agency's (EPA's) Office of Water (OW), Engineering and Analysis
Division (EAD).   This document was prepared under EPA Contract No.  68-C9-0019 by the
Environmental Services  Division of Viar & Company, a wholly-owned subsidiary of DynCorp.  The
methods were written by Interface, Inc. The American Petroleum Institute and its member companies
supported much of the methods development.
Disclaimer
      The methods in this compendium have been reviewed  and approved for publication by the
Engineering and Analysis Division of the U.S. Environmental Protection Agency.  Mention of trade
names or commercial products does not constitute endorsement or recommendation for use.

-------
                                                                          Introduction
       The U.S. Environmental Protection Agency (EPA) is promulgating effluent limitations guidelines
for the Offshore subcategory of the Oil and Gas Extraction Point-Source category at Subpart A of 40 CFR
Part 435, under the authority of the Clean Water Act.  The rule will in part prohibit diesel oil in drilling
fluids (muds)  and drill cuttings  from being discharged from offshore oil and  gas  platforms.   This
compendium of analytical methods supports the final rule.  The methods are being promulgated under
the authority of Sections 304(h), 501 (a),  and 308 of the Federal Water Pollution Control Act (FWPCA,
33 U.S.C. 1251, et.  seq.).    These sections require EPA to promulgate guidelines establishing test
procedures for the determination of pollutants in discharges and to require data collection whenever
necessary to carry out the objectives of the CWA.  The test procedures were developed by EPA's
Engineering and Analysis Division within the Office of Water's Office of Science and Technology.
       Method 1651 was proposed as part of the 40 CFR Part 435 rule (56 FR 10664-10715).  In Method
1651, oil  is removed  from  a mud sample using a thermal extraction apparatus called a "retort."  The
solution from  the retort is placed in  a separatory funnel and extracted using methylene chloride.  The
extract is  evaporated to dryness in a  Kuderna-Danish concentrator, and the weight of oil is determined
using an analytical balance.  The oil is then redissolved in methylene chloride, an internal standard is
added, and an aliquot is analyzed by gas chromatography (GC).  Diesel oil is identified by comparing
the pattern of  GC peaks with the pattern produced by reference diesel oil.
       The  American Petroleum Institute (API) and  its member companies  criticized Method  1651,
arguing that it was unable to reliably distinguish diesel oil from mineral oil and certain permitted mud
additives  and  that the retort apparatus used in Method 1651 could not be operated reproducibly.  In
response to this criticism, EPA worked with API to develop Methods 1662, 1654A, and 1663. A flow
chart of how these methods are employed is shown in Figure 1.  In the final rule, Methods 1662, 1654A,
and 1663  are being allowed as variants of Method 1651.
       Method 1662 is a laboratory extraction procedure employing a Soxhlet/Dean-Stark (SDS) extractor
for reliable removal of oil from the mud.  A total oil measurement by gravimetry (weighing) can be made
with this procedure, if desired.
       Methods 1654A and 1663 both use a portion of the extract from Method 1662.  A portion of the
extract from Method  1651  can be used  if desired, but  the SDS extraction procedure in Method  1662
provides improved precision and accuracy over the retort procedure in Method 1651.
       Method 1654A measures the polynuclear  aromatic hydrocarbon (PAH)  content of the  oil as
phenanthrene by high-performance liquid chromatography with an ultra-violet detector (HPLC/UV).  If
the PAH  content of the oil  is less than 0.35 weight percent,  the oil is mineral oil. If the PAH content
is equal to or greater than 0.35 weight percent, the oil is diesel oil or crude oil.
       Methods 1651  and 1663 both measure the presence and distribution of hydrocarbons in the oil by
gas chromatography with a flame ionization detector (GC/FID). Method 1651 uses pattern matching to
attempt to distinguish diesel oil from mineral or crude oil; Method 1663 uses the distribution of normal
aliphatic hydrocarbons (n-alkanes) in the Cg-Cjo range to distinguish diesel oil from crude oil.
       Using Method  1663, the presence of n-alkanes in  the Cg-C^ range indicates the presence of diesel
oil or  crude oil.   If less than 10 n-alkanes are present in the Cg-C^ range (at a signal-to-noise ratio
of 3 or greater for each n-alkane), diesel oil is not present.  If 10 or more n-alkanes  are present in the
C9-C24 range,  the percentage of n-alkanes in the C^-Cy, range are used to determine  if the oil is crude
oil.  The  oil is crude  oil if the C^-C-x, n-alkane content is greater than  1.2 percent of the total Cg-Cy,
n-alkane content.

-------
Introduction
       Methods 1651, 1654A, 1662, and 1663 all employ standardized 304(h) quality control.  This QC
requires instrument calibration and periodic verification, initial and ongoing demonstration of laboratory
capability, and recovery of spikes in samples to demonstrate that the method is applicable to each sample
type tested.
       Questions about the methods or this document should be addressed to:

              William A. Telliard, Director
              Analytical Methods Staff
              Engineering and Analysis Division (WH-552)
              U.S. Environmental Protection Agency
              401 M Street, S.W.
              Washington, DC 20460

               (202) 260-7531

-------
                                                                      Introduction
                                Method 1662
                                SDS extraction
                        Rotovap to approx imately 1 ml
                    Adjust volume to 5.0 ml with acetonitrile
               Evaporate 4.0 ml to dryness with nitrogen blowdown
                Determine total oil in 4.0-mL portion  by gravimetry
                                 Method 1654 A
               Determine PAH content of 1.0-mL portion by HPLC/UV
                    if PAH content <0.35 wt %, oi! is mineral oil
               If PAH content >0.35 wt %, oil may be diesel or crude oil
                                 Method 1663
              Determine n-alkane pattern of 1.0-mL portion by GC/FID
           If <10 n-alkanes present in Cg-C24 range @ S/N >3, no diesei
          If n-alkanes present in 09-630 range & ^25^30 n-a!kanes >1.2
                    % of total Cg-C3o n-alkanes, oil is crude oil
                                                                     CS2-10-S
Figure 1.     Differentiation of Diesel, Mineral, and Crude Oils by SDS Extraction,
            HPLC/UV, and GC/FID,  using Methods 1662, 1 654A, and 1663, respectively

-------
                                                              Contents
Introduction	  iii

Method 1651, Rev. A
      Total Oil and Diesel Oil in Drilling Muds and Drill Cuttings
      by Retort, Gravimetry, and GC/FID	  1

Method 1654, Rev. A
      PAH Content of Oil by HPLC/UV  	  23

Method 1662
      Total Extractable Material in Drilling Mud by SDS Extraction and Gravimetry ...  39

Method 1663
      Differentiation of Diesel and Crude Oil by GC/FID	  55

-------
Method 1651, Revision A
      Total Oil and Diesel Oil in
 Drilling Muds and Drill Cuttings
                   by Retort,
       Gravimetry, and GC/FID

-------
                                1651,  Revision  A
        Total Oil and Diesel Oil in Drilling Muds and Drill Cuttings
                       by Retort,  Gravimetry,  and GC/FID


1.     SCOPE AND APPLICA TION

1.1    This method is used to determine the total oil content and the identity and concentration of
       diesel oil in drilling mud samples. It is applicable to all mud types and may also be used to
       determine the total oil and diesel oil content of drill cuttings.
1.2    This method may be used for compliance monitoring purposes as part of the "Effluent
       Limitations Guidelines and New Source Performance Standards for the Offshore Subcategory
       of the Oil and Gas Extraction Point Source Category" [50 FR 34592].
1.3    When this  method is used to analyze samples for which there is no reference diesel oil, diesel
       oil identification shall be supported by at least one additional qualitative technique.
       Method 1625 provides gas chromatograph/mass spectrometer (GC/MS) conditions appropriate
       for the qualitative and quantitative confirmation of the presence of the components of diesel oil
       (References 1 and 2).
1.4    The detection limit of this  method is usually dependent upon the presence of other oils in the
       sample.  Excluding interferences, estimated detection limits of 200 mg/kg of total oil and
       100 mg/kg of diesel oil can be obtained.
1.5    Any modification of this method beyond those expressly permitted shall be considered as a
       major modification subject to application and approval of alternative test procedures under
       40 CFR 136.4 and 136.5.
1.6    The gas chromatography portions of this method are restricted to use by or under the
       supervision of analysts experienced in the use of gas chromatography and in the interpretation
       of gas chromatograms.  Each laboratory that uses this method  must generate acceptable results
       using the procedures described in Sections 8.2 and 12 of this method.

2.     SUMMARY OF METHOD

2.1    A weighed amount of drilling mud is distilled using a retort apparatus.  The distillate is
       extracted with methylene chloride and the extract is dried by passage through sodium sulfate.
       The extract is evaporated to dryness, and the total amount of oil in the sample is determined
       gravimetrically.  The oil, together with a measured volume of internal standard, is redissolved
       in a known volume of methylene  chloride, and an aliquot is injected into a gas chromatograph
       (GC).  The components  of the oil are separated  by the GC and detected using a flame
       ionization detector (FID).
2.2    Identification of diesel oil (qualitative analysis) is performed by comparing the pattern of GC
       peaks (retention times and  intensities) from the sample extract  with the pattern of GC peaks
       from a reference diesel oil  sample.  Identification of diesel oil  is established when the
       reference diesel and sample patterns agree per the criteria in this method.
2.3    Quantitative analysis of diesel oil  is performed using an internal standard technique.

-------
Method 1651A
3.     CONTAMINATION AND INTERFERENCES

3.1    Solvents, reagents, glassware, and other sample processing hardware may yield artifacts and/or
       elevated baselines causing misinterpretation of chromatograms.
       3.1.1   All materials shall be demonstrated to be free from interferences under the conditions
               of the analysis by running method blanks initially and with a set of samples.  Specific
               selection of reagents and purification of solvents by distillation in all-glass systems
               may be required.
       3.1.2   Glassware and, where possible, reagents are cleaned by solvent rinse or baking at
               450°C for a minimum of 1 hour.
3.2    There is no standard diesel oil.  Oil components, as seen by GC/FID, will differ depending
       upon the oil source, the production date, production process, and the producer.  In addition,
       there are three basic types of diesel oils:  ASTM Designations No. 1-D, No. 2-D, and
       No. 4-D.  The No. 2-D is the most common diesel oil; however, No. 2-D is sometimes
       blended with No. 1-D,  which  has  a lower boiling range.  For rigorous identification and
       quantification of diesel  oil in a drilling fluid sample by GC/FID, the chromatographic pattern
       from the diesel oil should be matched with the chromatographic pattern from a reference
       standard of the same diesel oil suspected to be in the sample.
3.3    To aid in the identification of interferences, the chromatographic pattern from a reference
       sample of drilling fluid prior to use is compared to the chromatographic pattern of the drilling
       fluid after use.  An interference is present when the pattern of the background oil  does not
       match, but contributes substantially to, the pattern of the diesel oil in the sample.
3.4    Mineral oils are often added to drilling  fluids for lubricity. These oils, when examined by
       GC/FID, contain some components common to diesel oil but have chromatographic patterns
       that are distinctly different from diesel oil. The analyst must first determine if the sample
       chromatogram shows the  presence of diesel, mineral, or a combination of both before reliable
       quantification can be performed. This method permits selection of GC peaks unique to diesel
       oil for determination of diesel oil in the presence of mineral  oil.

4.     SAFETY

4.1    The toxicity or carcinogenicity of each reagent used in this method has not been defined.
       Therefore, each chemical compound should be treated as a potential health hazard. From this
       viewpoint, exposure to these chemicals  must be reduced to the lowest possible level by
       whatever means  available.
4.2    The laboratory is responsible for maintaining a current awareness file of OSHA regulations
       regarding  the safe handling of the chemicals specified in this method.  A reference file of
       material handling data sheets (MSDSs) should also be made  available to all personnel involved
       in the chemical analysis.  Additional references to laboratory safety are available and have
       been identified (References 3 through 5) for the information of the analyst.
4.3    Methylene chloride has been classified as a known health hazard.  All steps in this method
       which  involve exposure to this compound shall be performed in an OSHA-approved fume
       hood.

-------
                                                                                 Method 1651A
5.    APPARA TUS AND MA TER/ALS
      NOTE: Brand names, suppliers, and pan numbers are for illustrative purposes only.
      No endorsement is implied.  Equivalent performance may be achieved using apparatus
      and materials other than those specified here, but demonstration of equivalent
      performance meeting the requirements of this method is the responsibility of the
      laboratory.

5.1   Sample bottles for discrete sampling.
      5.1.1  Bottle: 4-oz Boston round wide-mouth jar with PTFE-lined screw-cap (Sargent Welsh
              S-9184-72CA, or equivalent). New bottles are used as received with no further
              cleaning required.
      5.1.2  Bottle mailer: To fit bottles above (Sargent-Welsh 2306, or  equivalent).
5.2   Distillation apparatus.
      5.2.1  Retort: 20-mL retort apparatus  (IMCO Services Model No.  R2100 or equivalent).
      5.2.2  Glass wool:  Pyrex  (Corning 3950, or equivalent). Solvent-extracted or baked at
              450°C for a minimum of 1  hour.
5.3   Extraction/drying apparatus.
      5.3.1  Separatory funnel:  60-mL with PTFE stopcock.
      5.3.2  Drying column: Pyrex chromatographic column, 400 mm long by  15 to 20 mm i.d.,
              equipped with coarse-glass frit or glass-wool plug.
      5.3.3  Glass filtering funnel:  Crucible holder (Corning No. 9480, or  equivalent).
      5.3.4  Spatulas: Stainless  steel or PTFE.
5.4   Evaporation/concentration apparatus.
      5.4.1  Kuderna-Danish (K-D) apparatus.
              5.4.1.1   Evaporation flask: 500-mL (Kontes K-570001-0500, or equivalent),
                         attached  to concentrator tube with springs (Kontes K-662750-0012).
              5.4.1.2   Concentrator tube:  10-mL, graduated (Kontes K-570050-1025, or
                         equivalent) with calibration verified.  Ground-glass  stopper (size 19/22
                         joint) is  used to  prevent evaporation of extracts.
              5.4.1.3   Snyder column:  Three-ball macro  (Kontes K-503000-0232, or equivalent).
              5.4.1.4   Snyder column:  Two-ball micro (Kontes K-469002-0219, or equivalent).
              5.4.1.5   Boiling chips.
                         5.4.1.5.1 Glass or  silicon carbide:  Approximately 10/40 mesh, extracted
                                   with methylene chloride and baked at 450°C for a minimum of
                                   1 hour.
                         5.4.1.5.2 PTFE (optional):  Extracted with methylene chloride.
      5.4.2  Water bath:  Heated, with concentric ring cover, capable of temperature control
              (+2°C), installed in a fume hood.

-------
Method 1651A
       5.4.3   Sample vials: Amber glass, 1- to 5-mL with PTFE-lined screw- or crimp-cap, to fit
               GC autosampler.
5.5    Balances.
       5.5.1   Analytical:  Capable of weighing 0.1 mg. Calibration must be verified with class S
               weights each day of use.
       5.5.2   Top-loading:  Capable of weighing 10 mg.
5.6    Gas chromatograph (GC):  Analytical system with split injection, capillary column,
       temperature program with initial and final isothermal holds, and  all  required accessories,
       including syringes, analytical columns, gases, detector, and recorder.  The analytical system
       shall meet the performance specifications in Section 12.
       5.6.1   Column:  30 m  (±5 m) long by 0.25 mm (±0.02 mm)  i.d., 99% methyl, 1% vinyl,
               1.0 jan film thickness, bonded-phase fused-silica capillary (Supelco SPB-1, or
               equivalent).
       5.6.2   Detector:  Flame ionization.  This detector has proven effective in the analyses of
               drilling fluids for diesel oil and was used to develop the method performance
               statements in Section 16.  Guidelines for using alternative detectors are provided in
               Section 11.1.
5.7    GC data system:   Shall collect and record GC data, store GC runs in magnetic memory or on
       magnetic disk or tape, process GC data, compute peak areas, store calibration data including
       retention times and response factors, identify GC peaks through retention times, and compute
       concentrations.
       5.7.1   Data acquisition: GC data shall be collected continuously throughout the analysis and
               stored on a magnetic storage device.
       5.7.2   Response factors and calibration curves:  The data system shall be used to record and
               maintain lists of response factors and multi-point calibration curves (Section 7).
               Computations of relative standard deviation  (coefficient  of variation,  CV) are used  for
               testing calibration linearity. Statistics on initial (Section 8.2)  and ongoing
               (Section 12.5) performance shall be computed and maintained.
       5.7.3   Data processing: The data system shall search, locate, identify, and  quantify the
               compounds of interest in each GC analysis.  Software routines shall be employed to
               compute and record retention times and peak areas. Displays of chromatograms and
               library comparisons are required to verify results.

6.     REAGENTS

6.1    Sodium sulfate:  Anhydrous, ACS grade, granular.
6.2    Methylene chloride:  Nanograde or equivalent.
6.3    Reagent water:  Water in which the compounds of interest and interfering compounds are not
       detected by this method.
6.4    Internal standard: Dissolve 1.0 g of 1,3,5-trichlorobenzene (TCB, Kodak No. 1801 or
       equivalent) in 100 mL methylene chloride.   Store in glass and tightly cap with PTFE-lined lid

-------
                                                                                   Method 1651A
       to prevent loss of solvent by evaporation.  Label with the concentration and date.  Mark the
       level of the meniscus on the bottle to detect solvent loss.
6.5    Calibration standards: Calibration standards are prepared from the same diesel oil expected to
       be in the sample;  otherwise, No. 2 diesel oil is used.
       6.5.1   Prepare stock solutions of calibration standards at the concentrations shown in
               Table 1.  Weigh the appropriate amount of oil into a tared 10-mL volumetric flask
               and dilute to volume with methylene chloride.  Label each flask with the concentration
               and date.
       6.5.2   Using a micropipette or microsyringe, transfer  100 /xL of each reference standard
               solution (Section 6.5.1) to a GC injection vial.  Add 100 /*L of the TCB internal
               standard  (Section 6.4)  to  each vial and mix thoroughly. Calibration standards are
               made fresh daily to avoid solvent loss by evaporation.
6.6    QC  standard: Used for tests of initial (Section 8.2) and ongoing (Section 12.5) performance.
       Prepare a reference mud sample containing 20,000 mg/kg of diesel by adding 20.0 mg
       (±0.2 mg) of No. 2 diesel oil  and 10.0 g (+0.1 g) of EPA Generic Mud No. 8 to a clean
       retort cup (see Section 10.1).  Mix the mud and diesel oil thoroughly with a metal spatula.

7.     CALIBRATION

7.1    Establish gas chromatographic  operating conditions given  in Table 2. Verify that the GC
       meets the performance criteria  (Section 12) and  the estimated detection limit (Section 1.4).
       The gas chromatographic system is calibrated using the  internal standard technique.

       NOTE: Because each GC is slightly different, it may be necessary to adjust the
       operating conditions (carrier gas flow rate and column temperature and temperature
       program) slightly  until the retention times in  Table 3 are met.

7.2    Internal standard calibration procedure:  1,3,5-Trichlorobenzene (TCB) has been shown to be
       free of interferences from the diesel oils tested in the development of this method.  Check for
       acceptability by injecting 0.5 fjL of the internal standard solution (Section 6.4) into the
       GC/FID.  If a major peak other than the TCB peak appears in the chromatogram, interference
       with the peaks used for determination of diesel oil may occur.  In this case, the analyst must
       choose an alternative internal standard that is free from interferences.
       7.2.1   Inject 1 ^iL of each calibration standard containing the internal standard (Table  1 and
               Section 6.5.2) into the GC/FID.  The TCB will elute approx 8.5 minutes after
               injection.  For the GC/FID used in the  development of this method, the TCB internal
               standard peak was 30  to 50% of full scale at an attenuator setting of 8 x 10"n amp.
       7.2.2   Individual response factors.
               7.2.2.1   Tabulate the peak area responses against concentration for each  of the ten
                         largest peaks in the chromatogram (excluding the solvent peak, the internal
                         standard peak, and any peaks that elute prior to the internal standard peak).
                         (See Section 13 for guidance on peak selection.) Calculate response
                         factors (RF) for each peak using the following equation:

-------
Method 1651A
                                         Equation 1

                                        RF  (/
                   where:
                    As = Area of the peak to  be measured
                   C^ = Concentration of the internal standard, in mglkg
                   A^ = Area of the internal  standard peak
                    C  = Concentration of the peak to be measured, in mglkg
               7.2.2.2   If the RF is constant (< 15% CV) over the calibration range (Table 1), the
                         RF can be assumed to be invariant and the average RF can be used for
                         calculations. Alternatively, the results can be used to plot a calibration
                         curve of response ratios, AS/AU vs. RF.
               7.2.2.3   Calibration verification:  For each peak, the average RF or a point on the
                         calibration curve shall be verified on each working day by the
                         measurement of one or more calibration standards.  If the RF for any peak
                         varies from the RF obtained in the calibration by more than  +15%, the
                         test shall be repeated using a fresh calibration standard.  Alternatively, a
                         new calibration curve shall be prepared.
       7.2.3   Combined response factor:  To reduce the error associated with the measurement of a
               single peak, a combined response factor is used for computation of the diesel oil
               concentration. This combined response factor is the sum of the individual response
               factors as given  in Equations 2 or 3:


                                          Equation 2
                           RF Combined  = RF(l) +  RF(2)... + RF(n)
                Equation 3

RF Combined =
                                          [A'm + A* (2)"'+ A'
                where:
                         n = Number of individuals peaks
                Asm...A(rti = Areas of the individual peaks
                       C^ = Concentration  of the internal standard,  in mglkg
                       A^ = Area of the internal standard peak
                        Cs = Concentration  of the peak to be measured, in mglkg
8

-------
                                                                                   Method 1651A
8.     QUALITY ASSURANCE/QUALITY  CONTROL

8.1    Each laboratory that uses this method  is required to operate a formal quality assurance
       program (Reference 6).  The minimum requirements of this program consist of an initial
       demonstration of laboratory capability, an ongoing analysis of standards and blanks as a test of
       continued performance, analyses of spiked samples to assess accuracy, and analysis of
       duplicates  to assess precision.  Laboratory performance is compared to established
       performance criteria to determine if the results of analyses meet the performance
       characteristics of the method.
       8.1.1   The analyst shall make an initial demonstration of the ability to generate acceptable
               accuracy and  precision with this method. This ability is established as described in
               Section 8.2.
       8.1.2   The analyst is permitted to modify this method to improve separations or lower the
               costs of measurements, provided all performance requirements are met. Each time a
               modification is made to the method, the  analyst is required achieve the estimated
               detection limit (Section 1.4) and to repeat the procedure in Section 8.2 to demonstrate
               method performance.
       8.1.3   Analyses of spiked samples are required to demonstrate method accuracy.  The
               procedure and QC criteria for spiking are described in Section 8.3.
       8.1.4   Analyses of duplicate samples are required to demonstrate method precision.  The
               procedure and QC criteria for duplicates are described in Section 8.4.
       8.1.5   Analyses of blanks are required to demonstrate freedom from contamination.  The
               procedures and criteria for analysis of a  blank are described in Section 8.5.
       8.1.6   The laboratory shall, on an ongoing basis, demonstrate  through calibration verification
               and analysis of the QC standard (Section 6.6) that the analysis system is in control.
               These procedures are described in Section 12.
       8.1.7   The laboratory shall maintain records to define the  quality of data that is generated.
               Development of accuracy statements is described in Sections 8.3.4 and 12.5.
8.2   Initial precision and accuracy:  To establish the ability to generate acceptable precision and
       accuracy, the analyst shall perform the following operations.
       8.2.1   Retort, extract, concentrate, and analyze four samples of the QC  standard
               (Sections 6.6 and 10.1.3) according to the procedure beginning in Section 10.
       8.2.2   Using results of the set of four analyses,  compute the average recovery  (X) in mg/kg
               and the standard deviation of  the recovery (s) in mg/kg  for each sample by the
               internal standard method (Sections 7.2 and 14.2).
       8.2.3   For each compound, compare s and X with the corresponding limits for initial
               precision and accuracy in Table 4. If s and X meet the acceptance criteria, system
               performance is acceptable and analysis of samples may  begin. If, however, s exceeds
               the precision limit or X falls outside the range for accuracy, system performance  is
               unacceptable.  In this event, review this  method and the retort manufacturer's
               instructions, correct the problem,  and repeat the test.

-------
Method 1651A
8.3    Method accuracy:  The laboratory shall spike a minimum of 20%  (one sample in each set of
       five samples) of all drilling fluid samples.  This sample shall be spiked with the diesel oil that
       was added to the drilling fluid.  If a reference standard of diesel oil that was added to the
       drilling fluid is not available, No. 2 diesel oil shall be used for this spike.  If doubt of the
       identity and concentration of diesel oil in any of the remaining 80%  of the samples exists, that
       sample shall be spiked to confirm the identity and establish the diesel oil concentration.
       8.3.1   The concentration of the spike in the sample shall be determined as follows.
               8.3.1.1   If, as in compliance monitoring, the concentration of the oil in the sample
                         is being checked against a regulatory concentration limit, the spike shall be
                         at that limit or at 1 to 5 times higher than the background concentration
                         determined in Section 8.3.2,  whichever concentration is higher.
               8.3.1.2  If the concentration of the oil in a sample is not being checked against a
                         limit, the spike shall be at the concentration of the QC standard
                         (Section 6.6) or at 1 to 5 times higher than the background concentration,
                         whichever concentration is higher.
       8.3.2   Analyze one sample aliquot to determine the background  concentration (B) of total oil
               and of diesel oil.  If necessary, prepare a standard solution appropriate to produce  a
               level in the sample at the regulatory concentration limit or at 1 to 5 times the
               background concentration (per Section 8.3.1).  Spike a second sample aliquot with the
               standard solution and analyze it to determine the concentration after spiking  (A) of
               each analyte. Calculate the percent recovery  (P) of total  oil and of diesel oil using
               Equation 4:


                                          Equation 4

                                         p =  IQO(A-B)
                                                  T
                     where:
                     A = Concentration of analyte after spiking
                     B = Background  concentration of total oil  and diesel oil
                     T = True value of the spike
       8.3.3   Compare the percent recovery for total oil and for diesel oil with the corresponding
               QC acceptance criteria in Table 4.  If the results of the spike fail the acceptance
               criteria, and the recovery of QC standard in the ongoing precision and recovery test
               (Sections 10.1.3  and 12.5) is within the acceptance  criteria in Table 4, an interference
               may be present (see Sections 3 and 15 for identification of interferences).  In this case,
               the result may not be reported for regulatory compliance purposes.  If, however, the
               results of both the spike and the ongoing precision and recovery test fail the
               acceptance criteria, the analytical system is judged to be out of control and the
               problem must be immediately identified and corrected and the sample set reanalyzed.
       8.3.4   As part of the QA program for the laboratory, method accuracy for samples shall be
               assessed and records shall be maintained.  After the analysis of five spiked samples in
               which the recovery passes the test in Section 8.3, compute the average percent

10

-------
                                                                                  Method 1651A
               recovery (P) and the standard deviation of the percent recovery (sp).  Express the
               accuracy assessment as a percent recovery interval from P  2sp to P +  2sp. For
               example, if P = 90%  and  sp =  10% for five analyses of diesel oil, the  accuracy
               interval  is expressed as 70 to  110%.  Update the accuracy assessment on a regular
               basis (e.g., after each  five to ten new accuracy measurements).
8.4    The laboratory shall analyze duplicate  samples for each drilling fluid type at a minimum of
       20% (one sample for each set of five samples).  A duplicate sample shall consist of a
       well-mixed, representative aliquot of the sample.
       8.4.1   Analyze one sample in the set in duplicate per the procedure beginning  in Section 10.
       8.4.2   Compute the relative percent difference (RPD) between the two results per the
               following equation:


                                          Equation 5

                                    RPD = —Lj	iL x 100
                                          (D, + D2)/2

                where'.
                Dj  = Concentration of diesel oil in the sample
                D2  = Concentration of diesel oil in the second (duplicate)  sample
       8.4.3   The relative percent difference for duplicates shall meet the acceptance criteria in
               Table 5.  If the criteria are not met, the analytical system shall be judged to be out of
               control, and the problem must be immediately identified and corrected and the sample
               set reanalyzed.
8.5   Blanks:  Reagent water blanks are analyzed  to demonstrate freedom from contamination.
       8.5.1   Extract and concentrate a reagent water blank initially and with each sample set
               (samples started through the analysis on the same day,  to a maximum of five
               samples).  Analyze the blank immediately after analysis of the QC standard
               (Section 6.6) to demonstrate freedom from contamination.
       8.5.2   If any of the components of diesel oil or any potentially interfering compound is
               detected in a blank,  analysis of samples is halted until the source of contamination is
               eliminated and a blank shows  no evidence of contamination.
8.6   Compare the concentration of the total oil (Section 14.1.2) determined  gravimetrically with the
       diesel oil concentration determined  by  GC/FID (Section 14.2.2).  If the diesel oil concentration
       exceeds the gravimetric oil concentration, the analysis has been performed improperly.
       Correct the error or repeat the sample  analysis beginning with Section  10.
8.7   The specifications contained  in this method can be met if the apparatus used is calibrated
       properly, then maintained in  a calibrated state. The standards used for  initial precision and
       recovery (IPR, Section 8.2) and ongoing precision and recovery (OPR, Section 12.5) should
       be identical, so that the most precise results will be obtained. The GC instrument  will provide
                                                                                              11

-------
Method 1651A
      the most reproducible results if dedicated to the settings and conditions required for the
      analyses given in this method.
8.8   Depending on specific program requirements, field replicates and field spikes ofdiesel oil into
      samples may be required to  assess the precision and accuracy of the sampling and sample
      transportation techniques.

9.    SAMPLE COLLECTION, PRESERVATION, AND HANDLING

9.1   Collect drilling fluid samples in wide-mouth glass containers following conventional sampling
      practices (Reference 7).
9.2   Samples must be representative of the entire bulk drilling fluid. In some instances, composite
      samples may be required.
9.3   Maintain samples at 0 to 4°C from the time of collection until extraction.
9.4   Sample and extract holding times  for this method have not yet been established.  However,
      based on tests of wastewater for the analytes determined  in this method, samples shall be
      extracted within seven days of collection and extracts shall be analyzed within 40 days of
      extraction.
9.5   As a precaution  against analyte and solvent loss or degradation, sample extracts are stored in
      glass bottles with PTFE-lined caps, in the dark, at -20 to -!0°C.

10.   SAMPLE EXTRACTION AND CONCENTRATION

10.1  Retort.
      10.1.1 Tare the retort sample cup and cap to the nearest 0.1 g.  Transfer a well-homogenized
              and representative portion of the drilling fluid to be tested into the sample cup.  Do
              not fill the retort cup to the top so that excess sample must be wiped off.  Place the
              cap on the cup and  reweigh.  Record the weight of the sample to the nearest 0.1 g.

      NOTE: On agitation, most drilling fluids entrain air as small bubbles.  The extent of
      air entrainment is uncertain  and is difficult to detect when the mud is poured into the
      retort cup. By weighing  the drilling fluid, the quantitative detection ofdiesel oil is
      improved.  In addition, by using a gravimetric measurement of the amount of sample,
      the retort cup does not need to be completely filled.  This procedure avoids the error
      that occurs when the cup is filled  and the oil rises to the  surface of the sample and must
      be wiped off (as occurs if the manufacturer's instructions are followed), thus resulting
      in a loss of oil.

      10.1.2 Follow  the manufacturer's instructions for retort of the drilling fluid.  Substitute 6 g
              of loosely packed glass wool for the steel wool in the manufacturer's instructions and
              distill the sample into a glass receiver.  The presence of solids in the distillate require
              that the distillation be rerun  starting with  a new portion of sample,  Placing more
              glass wool in the retort expansion chamber, per  the manufacturer's  instructions, will
              help prevent the solids from being carried over in the distillation.
12

-------
                                                                                  Method 1651A
       10.1.3 QC standard: Used for tests of initial (Section 8.2) and ongoing (Section 12.5)
               precision and accuracy.  For the initial set of four samples (Section 8.2) and for each
               set of samples started through the retort process on the same working day (to a
               maximum of five), prepare a QC sample as follows.
               10.1.3.1  Place the QC standard (Section 6.6) in the retort cup beginning in
                         Section 10.1.
               10.1.3.2  Analyze the QC standard beginning with Section 10.1.2 then proceeding to
                         Section 10.2.
       10.1.4 Blank:  For the  initial set of four samples (Section 8.2) and for each set of samples
               started through the retort process on the same working day (to a maximum of five),
               prepare a blank  as follows.
               10.1.4.1     Place 10 mL of reagent water in a clean, tared retort cup and weigh to
                            the nearest mg.  Record the weight of the reagent water.
               10.1.4.2     Analyze the blank beginning with Section 10.1.2 then proceeding to
                            Section 10.2.
10.2  Extraction and drying.
       10.2.1  After the distillation is complete, pour the retort distillate into a 60-mL  separatory
               funnel.   Quantitatively rinse the inner surfaces of the retort stem and  condenser with
               methylene chloride into the separatory funnel.  Rinse the receiver with two full
               receiver volumes of methylene chloride and add to the separatory funnel.
       10.2.2 Stopper and shake the funnel for one minute, with periodic venting to prevent a
               buildup of gas pressure.  Allow the layers to separate.
       10.2.2 Prepare a glass filtering funnel by plugging the bottom with a piece of glass wool and
               pouring in 1  to 2 inches  of anhydrous sodium sulfate.  Alternatively,  a drying column
               may be used. Wet the funnel or column with a small  portion of methylene chloride
               and allow the methylene chloride to drain to a waste container.
       10.2.3 Place the glass filtering funnel or drying  column  into the top of a Kuderna-Danish
               (K-D) flask equipped with a preweighed  10-mL receiving flask.  Add a preweighed
               boiling chip to the receiving flask.   Drain the methylene chloride (lower) layer into
               the glass filtering funnel or drying column, and collect the extract in the K-D flask.
       10.2.4 Repeat the methylene chloride extraction twice more, rinsing the receiver with two
               thorough washings each time and draining each methylene chloride extract through the
               funnel or drying column into the K-D flask.
10.3  Concentration.
       10.3.1  Place a Snyder column on the K-D flask.  Prewet the Snyder column by adding about
               1 mL of methylene chloride to the  top.  Place the K-D apparatus on a hot water bath
               (60 to 65°C) so  that the concentrator tube is partially immersed in the hot water and
               the entire lower  rounded surface of the flask is bathed with hot vapor.  Adjust the
               vertical  position and the water temperature as required to complete the concentration
               in 15 to 20 minutes. At the proper rate of distillation, the balls of the column will
               actively chatter but the chambers will not flood with condensed solvent.  Concentrate
                                                                                             13

-------
Method 1651A
              the sample until it is free of methylene chloride.  Remove the K-D apparatus from the
              hot water bath and allow to cool.
       10.3.2 Weigh and record the final weight of the receiving flask.
       10.3.3 Dissolve the oil in methylene chloride and adjust the final volume to 0.40 mL
              (400 /xL).  If the extract did not concentrate to a final volume of 0.40 mL or less,
              adjust the final  volume to 4.0 mL.

 11.   GAS CHROMATOGRAPHY

 11.1  Table 3 lists the retention times that can be achieved under the conditions in Table 2 for the
       n-alkanes of interest.  The chromatogram in Figure 1 can be used as guidance in peak selection
       when peaks other than the n-alkanes are used.  Other retort devices, columns, chromatographic
       conditions, or detectors  may be used if the estimated detection limits (Section 1.4) and the
       requirements of Section 8.2 are met.
 11.2  Using a micropipette or microsyringe, transfer equal 100-/iL volumes of the sample extract or
       QC standard extract (Section 10.3.3) and the TCB internal standard solution (Section 6.4) into
       a GC injection vial.  Cap tightly and mix thoroughly.
 11.3  Inject 1 [iL of the sample extract or reference standard into the GC using the conditions in
       Table 2.
 11.4  Begin data collection and the temperature program at the time of injection.
 11.5  If the area of any peak exceeds the calibration  range of the system, make a ten-fold  dilution of
       the extract (Section 10.3.3), mix a 100-/*L aliquot of this dilute extract with 90 /iL of the
       internal standard solution (Section 6.4), and reanalyze.

 12.   SYSTEM AND LABORATORY PERFORMANCE

 12.1  At the beginning of each 8-hour shift during which analyses are  performed,  GC calibration is
       verified.  For these tests, analysis of a calibration standard prepared from the 300-mg/mL
       reference standard (Section 6.5.2) shall be used to verify all performance criteria.  Adjustment
       and/or recalibration (per Section 7) shall be performed until all performance criteria are met.
       Only after all performance criteria are met may the QC standard, blank, and samples be
       analyzed.
 12.2  Retention times.
       12.2.1 Retention time  of the  internal standard:  The absolute retention time of the  TCB
              internal standard shall be within the range of 7.96 to 8.08 minutes.
       12.2.2 Relative retention times of the n-alkanes: The retention times  of the n-alkanes relative
              to the TCB internal standard shall be within the limits given in Table 3.
 12.3  Calibration verification.
       12.3.1 Compute the response factor for each  peak by the internal standard technique
              (Section 7.2).
       12.3.2 For each peak, compare the response factor with the response factor from the initial
              calibration (Section 7.2.2). If all  response factors are within ± 15% of their
              respective values in the initial calibration, system calibration has been  verified. If

 14

-------
                                                                                   Method 16 51A
               not, prepare a fresh calibration standard and repeat the test (Section 12.1), or
               recalibrate (Section 7).
12.4  Multiple GC peaks:  Each peak selected for use in the calculation shall be a single, distinct GC
       peak.
12.5  Ongoing precision and accuracy.
       12.5.1 Compute the total oil concentration and the concentration of diesel oil in the QC
               standard in each sample set (Section  10.1.3) prior to analysis of any sample  in the set.
       12.5.2 Compare the concentration with the QC limit in Table 4.  If the concentrations of total
               oil and of diesel oil in the QC standard meet the acceptance criteria, system
               performance  is acceptable and analysis of samples  may proceed.  If, however, the
               concentrations do not meet the acceptance criteria, system performance  is  unac-
               ceptable.  In  this event, correct the problem, reprocess the sample set (Section 10),
               and repeat the ongoing precision and accuracy test (Sections 10.1.3 and  12.5).
       12.5.3 Add results that pass the specifications in  Section 12.5.2 to initial and previous
               ongoing data. Update QC charts to form  a graphic representation of continued
               laboratory performance.  Develop  statements of laboratory accuracy for total oil and
               diesel oil in drilling fluids by calculating the average percent recovery (R) and the
               standard deviation of percent recovery (sr). Express  the accuracy statement as a
               recovery interval from R  2sr to R + 2sr.  For example, if R  = 95% and sr = 5%,
               the accuracy  is 85 to 105%.

13.   QUALITA TIVE DETERMINA TION
If less than ten of the GC peaks used to calculate the combined RF (Section 7.2.2) are present in the
sample and the QC tests (Sections 8 and 12) for the sample set are acceptable, diesel oil is not present
in the sample.  If the  ten GC peaks used to calculate the combined RF are present in the analysis of
the  sample,  diesel oil  is identified by comparing the relative retention times and relative areas of
peaks to the respective peaks in the diesel oil calibration standard.  If either the relative retention
times or the relative areas disagree and  the QC tests  (Sections  8 and  12) are acceptable, then an
interference is suspected,  and Method 1625 (Revision C or greater) shall be used  to determine the
presence of the components of diesel oil in the sample.
13.1  Relative retention times:  If the n-alkane peaks are used for identification of diesel  oil, the
       relative retention times shall agree within the limits in Table 3.  If peaks other than the
       n-alkanes are used, the relative retention times shall agree within +2% of the retention times
       of the peaks used for calibration (Section 7).
13.2  Distribution of peak area ratios  (Reference  8): Diesel oil is further identified by comparing
       the distribution of the area ratios of ten peaks in  the chromatogram of the  diesel  oil calibration
       standard (Section 6.5) to these same ratios in  the chromatogram of the sample  extract.
       13.2.1 Using the largest peak of the ten as reference,  divide the area of each of the other
               nine peaks by the area of this largest peak. Repeat this division process for  the  same
               ten peaks  in the calibration standard.
       13.2.2 If all ratios agree within ±21%, diesel has been positively identified.  The quantity of
               diesel oil is then determined per Section 14.
                                                                                               15

-------
Method 1651A
14.   QUANTITATIVE DETERMINATION

14.1   Total oil by gravimetry.
       14.1.1 Subtract the weight of the preweighed receiving flask and boiling chip
              (Section 10.2.3) from the final weight of the receiving flask (Section 10.3.2).
       14.1.2 Calculate the concentration of total oil in the sample using the following equation:
                                         Equation 6
                                                W,
                                   C (mg/kg)  =  _/ x 1000

                  where;
                  W, = Final  weight of oil, in  mg (from Section 14.1.1)
                  Ws - Wet weight of sample,  in grams  (from Section 10.1.1)
14.2  Diesel oil by gas chromatography.
       14.2.1 Compute the concentration of diesel oil in the sample extract using the combined
              response factor given in Section 7.3.3 for the ten largest peaks chosen in Section 13
              using the following equation:


                                         Equation 7

                            Ccx (mg/mL) =
                                            (A^RF Combined)

                     where:
                     Cex = Concentration of the oil in the extract
                     ^ = Concentration of the internal standard, in mg/mL
                     As = Area of the peak to be measured
                     A.a = Area of the internal standard peak

                     (For RF Combined, see Equations 2 and 3)
       14.2.2 Calculate the concentration of diesel oil (in mg/kg) in the sample as follows:


                                         Equation 8

                                C (mg/kg) = —-—— x 1000
                                                  j
               where:
               CM = Concentration of the oil in the extract
               Vex = Final extract volume, in mL (from Section 10.3.3 or 14.2.3)
               W = Wet weight of sample, in g (from Section 10.1.1)
16

-------
                                                                                  Method 1651A
       14.2.3 If area of any peak in the chromatographic pattern exceeds the calibration range of the
              GC, the extract is diluted by a factor of 10 with methylene chloride, 100 ^L is
              withdrawn and mixed with 90 pL of the internal standard solution (Section 6.4), and
              the diluted extract is reanalyzed.

14.3  Results of analyses of diesel oil in drilling fluids are reported in units of mg/kg (wet weight) to
       three significant figures.  Results for samples that have been diluted are reported at the least
       dilute level at which the peak areas are within the calibration range (Section 14.2.3).

7 5.   COMPLEX SA MPLES

15.1   The most common  interference in the determination of diesel oil are from mineral oil in
       drilling fluids, proprietary additives, and naturally occurring hydrocarbons from crude-oil-
       bearing formations.
15.2  The presence  of mineral oil or other interfering oils and additives can often be determined by
       comparing the pattern of chromatographic peaks in the sample with the patterns of
       chromatographic peaks in the reference standard (Sections  6.5 and 10.1.3) and in the spiked
       sample (Section 8.3).
15.3  In cases where there is a mixture of diesel and mineral oil, the analyst may have to choose
       some of the smaller early- or late-eluting peaks  present in the chromatographic pattern of the
       diesel oil, and not present in the chromatographic pattern of the mineral oil, to determine the
       diesel content. Quantification using these peaks is performed by using these peaks for
       calibration (Section 7) and for determination of  the final concentration (Section  14).
15.4  Method 1625  (Revision C or later) shall be used to determine the components of diesel oil if
       diesel oil  cannot be determined by this method in the presence of these interferences.  If this
       method and Method 1625 are both used  to determine the presence of the components of diesel
       oil, the results from the determination using Method 1625  take precedence over the results
       from the determination using this method.

16.   METHOD PERFORMANCE

16.1   This method was developed by two laboratories that tested for diesel oil in drilling fluids
       (mainly drilling muds) over a two-year period.  The performance data for this method are
       based on the performance of the method in these two laboratories (Reference 9).
16.2  The most commonly occurring drilling fluid in the tests of this method was a seawater
       lignosulfonate mud (EPA Generic Mud No.  8).  The  estimated detection limit for diesel oil in
       this mud is  100 mg/kg.
                                                                                             17

-------
Method 1651A
                                    References

1.    Brown, John S.  "Organic Chemical Characterization of Diesel and Mineral Oils Used as
      Drilling Mud Additives."  Proceedings of Tenth Annual Analytical Symposium, USEPA,
      Industrial Technology Division (WH-552), 401 M St., S.W., Washington, DC 20460:  March
      19-20, 1986.

2.    Brown, John S.  "Final Report for Research Program on: Organic Chemical Characterization
      of Diesel and Mineral Oils used as Drilling Mud Additives." Phase II, Contract Reference
      Agreement No. 501-P-5476R, to Offshore Operators Committee,  Environmental
      Subcommittee. Houston, TX.  (Prepared by Battelle Ocean Science and Technology
      Department, 397 Washington St., Duxbury, MA 02332.)

3.    "Carcinogens—Working With Carcinogens." Department of Health,  Education, and Welfare,
      Public Health Service, Centers for Disease Control [available through National Technical
      Information System, 5285 Port Royal Road, Springfield, VA 22161,  document no. PB-
      277256]: August 1977.

4.    "OSHA Safety and Health Standards, General Industry [29 CFR 1910], Revised."
      Occupational Safety and Health Administration, OSHA 2206.  Washington, DC: January 1976.

5.    "Safety in Academic Chemistry Laboratories (3rd Edition)."  American Chemical Society
      Publication, Committee on Chemical Safety. Washington, DC: 1979.

6.    "Handbook of Analytical Quality Control in Water and Wastewater Laboratories." USEPA,
      EMSL-Ci, EPA-600/4-79-019. Cincinnati, OH: March  1979.

7.    "Standard Practice for Sampling Water,"  ASTM Annual Book of Standards, Part 31, D3370-
      76, ASTM. Philadelphia, PA: 1980.

8.    Boehm, Paul D., and Sauer, Theodor C., "Review and Evaluation of Draft EPA Method 1651
      Total Oil and Diesel Oil in Drilling Discharges by Retort, Gravimetry, and GC-FID."  Battelle
      Ocean Sciences, 397 Washington St., Duxbury, MA 02332: March 4, 1988.

9.    Rushneck, D.R., and Eynon, B.P., "Precision and Recovery Analysis of DPMP Diesel
      Measurements." Memorandum to Dennis Ruddy, USEPA, Industrial Technology Division
      (WH-552), 401 M. St. S.W., Washington, DC 20460 (23 August 1987,  draft).
18

-------
                                                                           Method 1651A
Table 1.  Concentration of Calibration Standards

               Expected Concen-   Wt of Diesel oil in   Concentration in
               tration in Sample    10 mL volumetric*   standard solution
                    (mg/kg)               (g)               (mg/mL)

                     50,000         use undiluted oil
                     30,000                 7.6              760
                     10,000                 3.0              300
                      5,000                 1.5              150

              *Weigh oil to the nearest mg
Table  2.  Gas Chromatographic Operating Conditions*

Injection port, transfer line, and detector temperatures: 275 °C
Column temperature program:
   Initial temperature:  90°C
   Initial time: 0 minutes
   Ramp: 90-250°C @ 5°C per minute
   Final temperature: 250°C
   Final hold:  10 minutes, or until all peaks have eluted
Carrier gas and flow rates:
   Carrier:  Nitrogen or helium
   Velocity: 20-40 cm/sec @ 90 °C
   Split ratio:  0-120:1**
   Makeup gas:  As required by manufacturer
Hydrogen and air flow rates: As specified by manufacturer
Detector amplifier settings: 10""  amp full scale.  Attenuation is adjusted so that the highest
   peaks are on scale in the most concentrated standard.
Recorder: Chart speed of 1-2 cm/min (fixed).

*  Conditions are approximate and can be adjusted to meet the performance criteria in Section 12
   (see  the note in Section 7.1)
**    Lower split ratios may give more reproducible results
                                                                                    19

-------
Method 1651A
Table 3. Retention Times and Relative Retention Time Limits for Major
         Components of Diesel Oil
                                         Retention Time
                    Compound

                    TCB
                    n-C12
                    n-C13
                    n-C14
                    n-C1B
                    n-C16
                    n-C17
                    n-C18
                    n-C19
                    n-C20
                    n-C21
                    n-C22
                    n-C23
                    n-C24
Mean
(minutes)
8.0
9.9
12.6
15.3
17.9
20.4
22.9
25.2
27.3
29.4
31.5
33.4
35.3
37.1

Relative
1.00-1.00
1.22-1.24
1.55-1.57
1.89-1.92
2.21-2.25
2.52-2.56
2.82-2.88
3.12-3.15
3.39-3.43
3.66-3.71
3.90-3.97
4.14-4.21
4.37-4.45
4.58-4.69
Table 4. QC Acceptance Criteria for Precision and Recovery
     Analyte

 Total oil by grav

 Diesel oil by GC
Test concentration
     (mg/kg)

     20,000
       n*

     20,000
       n*
Limit for s
 (mg/kg)

  3,400
  0.17n

  3,600
  0.18n
 Range for X
   (mg/kg)

18,000-23,700
 0.88n-1.16n

17,200-20,300
 0.80n-1.08n
 Range for P
   (mg/kg)

16,700-24,900
 0.82n-1.22n

13,600-21,400
 0.73n-1.14n
 *  For other test concentrations in the range of 1,000-50,000 mg/kg, assuming a spike to
    background ratio of 5:1.
Table 5. QC Acceptance Criteria for Duplicates
                      Concentration
                         Detected
                         (mg/kg)

                            500
                            750
                          1,000
                          2,000
                          5,000
                         10,000
                         20,000
                         50,000
                     Relative % Difference
                    Total Oil

                     36
                     30
                     28
                     24
                     21
                     21
                     20
                     20
             Diesel Oil

                 94
                 68
                 54
                 34
                 22
                 18
                 16
                 15
20

-------
                                                                              Method 1651A
  Diesel Oil in Drilling Mud




  Reference Diesel Oil


  Mineral Oil
                                                                                      A52-010-4
Figure 1.  Gas Chromatography of Diesel Oil in Drilling Mud,  Reference Diesel Oil, and
          Mineral Oil
                                                                                       21

-------
Method 1654, Revision A
       PAH Content of Oil by
                 HPLC/UV

-------
                         Method  1654,  Revision A
                         PAH Content of Oil by HPLC/UV

1.    SCOPE AND APPLICA TION

1.1   This method is designed to determine the polynuclear aromatic hydrocarbon (PAH) content of
      oil by high-performance liquid chromatography (HPLC) with an ultra-violet absorption (UV)
      detector. The PAH content is measured and reported as phenanthrene.
1.2   This method is for use in the Environmental Protection Agency's (EPA's) survey and
      monitoring programs under the Federal Water Pollution Control Act.
1.3   For oil in drilling muds, this method is designed to be used in conjunction with the extraction
      procedure in EPA Method  1662.
1.4   The level of PAH in Table  1 typifies the minimum level that can be detected in oil with this
      method.
1.5   Any modification of this method beyond those expressly permitted shall be considered as a
      major modification subject  to application and approval of alternative test procedures under
      40 CFR 136.4 and 136.5.
1.6   This method is restricted to use by or under the supervision of analysts experienced in the use
      of HPLC systems and in the interpretation of liquid chromatograms. Each analyst must
      demonstrate the ability to generate acceptable results with this method using the procedure
      described in Section 8.2.

2.    SUMMARY OF METHOD

2.1   An oil sample is diluted in  acetonitrile and a 20-/*L aliquot is injected into the HPLC. The
      PAHs are partially separated by HPLC and detected with the UV detector.
2.2   Identification of PAH (qualitative analysis) is performed by comparing the response of the UV
      detector to  the response during the retention-time range characteristic of the PAH in diesel oil.
      PAH is present when a response occurs during this retention-time range.
2.3   Quantitative analysis is performed by calibrating the HPLC with phenanthrene using an
      external standard technique, and using the calibration factor to determine  the concentration of
      PAH in the sample.
2.4   Quality is assured through reproducible calibration and testing of the extraction and HPLC
      systems.

3.    INTERFERENCES

3.1   Solvents, reagents, glassware, and other sample processing hardware may lead to discrete
      artifacts and/or elevated baselines causing misinterpretation of chromatograms.
      3.1.1   All materials used  in the analysis shall be demonstrated to be free from interferences
              by running method blanks initially and with each sample batch (samples started
              through the extraction process at the same time, to a maximum often). Specific
              selection of reagents and purification of solvents by distillation in all-glass systems
              may be required.
                                                                                         25

-------
Method 1654A
       3.1.2  Glassware and, where possible, reagents are cleaned by solvent rinse and/or baking at
              450°C for a minimum of 1 hour.
3.2    When used in conjunction with Method 1662, blanks extracted in that method are treated as an
       integral part of this method.
3.3    Interferences co-extracted from samples may vary from source to source, depending on the
       diversity of the site being sampled.

4.     SAFETY

4.1    The toxicity or carcinogenicity of each compound or reagent used  in this method has not been
       precisely defined; however, each chemical should be treated as a potential health hazard.
       Exposure to these chemicals must be reduced to the lowest possible level.
4.2    The laboratory is responsible for maintaining a current awareness file of OSHA regulations
       regarding the safe handling of the chemicals specified  in this method. A reference file of
       material safety data sheets (MSDSs) should also be made available to all personnel involved in
       the chemical analysis.  Additional information on laboratory safety can be found in
       References 1 through 3.
4.3    Methylene  chloride has been classified as a known health hazard.  All steps in this method
       which involve exposure to this compound shall be performed in an OSHA-approved fume
       hood.

5.     APPARATUS AND MATERIALS

       NOTE: Brand names,  suppliers, and part numbers are for illustrative purposes only.
       No endorsement is implied. Equivalent performance may be achieved using apparatus
       and materials other than those specified here, but demonstration of equivalent
       performance meeting the requirements of this method is the responsibility of the
       laboratory.

5.1    Equipment for glassware cleaning.
       5.1.1  Laboratory sink with overhead fume hood.
       5.1.2  Kiln:   Capable of reaching 450°C within 2  hours and holding 450°C within + 10°C,
              with temperature controller and safety switch  (Cress Manufacturing Co, Sante Fe
              Springs, CA, B31H or X31TS, or equivalent).
5.2    Equipment for sample preparation.
       5.2.1  Laboratory fume hood.
       5.2.2  Analytical balance:  Capable of weighing 0.1  mg.
       5.2.3  Glassware.
              5.2.3.1  Disposable pipettes:  Pasteur, 150 mm long by 5 mm i.d. (Fisher Scientific
                        13-678-6A, or equivalent).
              5.2.3.2  Glass pipettes:  1.0- and  10-mL, accurate to 1% or  better.
              5.2.3.3  Volumetric flasks:  Glass, 10- and 100-mL.

26

-------
                                                                                 Method 1654A
       5.2.4   Sample vials: Amber glass, 2- to 5-mL with PTFE-lined screw-cap, to fit HPLC
               autosampler.
5.3    High-performance liquid chromatograph (HPLC): An analytical system complete with pumps,
       sample injector, column oven, and ultra-violet (UV) detector.
       5.3.1   Pumping system:  Capable of isocratic operation and producing a linear gradient from
               50% water/50% acetonitrile to 100% acetonitrile in 10 minutes (Waters 600E, or
               equivalent).
       5.3.2   Sample injector:  Capable of automated injection of up to 30 samples (Waters 700, or
               equivalent).
       5.3.3   Column oven:  Capable of operation at room ambient to  50°C (Waters TCM, or
               equivalent).
       5.3.4   Column:  Two C18 columns, 150 mm long by 4.6 mm i.d., 300 angstroms (Vydac
               201 TP5415, or equivalent) connected in series, preceded by one  C18 guard column,
               30 mm long by 4.6 mm i.d., 300 angstroms (Vydac 201  GCC54T, or equivalent),
               operated at the conditions shown in Table 1.
       5.3.5   Detector:  UV operated at 254 nm (Waters 490E, or equivalent).
5.4    Data system.
       5.4.1   Data acquisition:  The data system shall collect and record LC peak areas and
               retention times on magnetic media.
       5.4.2   Calibration:  The data system shall be used to calculate and maintain lists of
               calibration factors (response divided by concentration) and multi-point calibration
               curves. Computations of relative standard deviation (coefficient of variation) are used
               to test calibration linearity.
       5.4.3   Data processing: The data system shall be used to search, locate, identify,  and
               quantify the compounds of interest in each analysis.  Displays of chromatograms are
               required to verify results.
       5.4.4   Statistics on initial (Section 8.2) and ongoing (Section 12.6)  performance  shall be
               computed and maintained.

6.     REAGENTS

6.1    Solvents.
       6.1.1   Sample preparation:  Methylene chloride, distilled in glass (Burdick and Jackson, or
               equivalent).
       6.1.2   HPLC: Methanol, acetonitrile, and water, HPLC quality.
6.2    Standards: Purchased as solutions or mixtures with certification to their purity, concentration,
       and authenticity, or prepared  from materials of known purity and  composition.  If compound
       purity is 96% or greater, the  weight may be used without correction to compute the
       concentration of the standard. If PAH in oil from drilling mud is to  be tested, the diesel  oil
       standard used in this method  should be from the oil used on the drilling rig from which the
       mud sample is taken.  If this  oil  is not available,  No. 2 diesel oil from a local source may be
       substituted.

                                                                                            27

-------
Method 1654A
       6.2.1   Stock solutions:  Prepare in methanol or acetonitrile for injection into the HPLC.
               Observe the safety precautions in Section 4.
               6.2.1.1   Diesel oil (1.25 mg/mL):  If QC extracts from Method 1662 are to be
                         tested, use the oil that was spiked to produce these extracts.  Weigh
                         125 mg of diesel oil to three significant figures in a 100-mL ground-glass-
                         stoppered volumetric flask and fill to the mark with acetonitrile.  After the
                         oil is  completely dissolved, transfer the solution to a clean 150-mL bottle
                         with PTFE-lined cap.
               6.2.1.2   Polynuclear aromatic hydrocarbons—naphthalene, phenanthrene, and
                         indeno[l,2,3-o/]pyrene:  Dissolve an appropriate amount of reference
                         material in a suitable solvent. For example, weigh 10.0 mg of naphthalene
                         in a 10-mL volumetric flask and fill to the mark with methanol. After the
                         naphthalene is completely dissolved, transfer the solution to a 15-mL vial
                         with PTFE-lined cap.
               6.2.1.3   Stock solutions should be checked for signs of degradation prior to the
                         preparation of calibration or performance test standards.
       6.2.2   PAH calibration  standards (CAL):  Dilute and mix the stock solutions
               (Section 6.2.1.2) in acetonitrile to produce the calibration standards shown  in Table 2.
               The three solutions permit the response of phenanthrene to be measured as  a function
               of concentration, and naphthalene and indeno[l,2,3-o/]pyrene permit the retention-
               time window for  PAH to be defined. The medium-level solution is used for
               calibration verification (Section 12.2).
       6.2.3   Precision and recovery  standard:  The stock solution of diesel oil (Section 6.2.1.1) is
               used for initial precision and recovery (IPR; Section 8.2) and ongoing precision and
               recovery (OPR; Section 12.6).
       6.2.4   Stability of solutions.
               6.2.4.1   When not being used, standards are stored in the dark at -20 to  -10°C in
                         screw-capped vials with PTFE-lined lids.  A mark is placed on the vial at
                         the level  of the solution so that solvent loss by evaporation can be
                         detected.  The vial is brought to room temperature prior to use. Any
                         precipitate is redissolved and solvent is  added if solvent loss has occurred.
               6.2.4.2   Standard solutions used for quantitative purposes (Sections 6.2.1 through
                         6.2.3) shall  be analyzed within 48 hours of preparation and on a monthly
                         basis thereafter for signs of degradation.  Standards will remain acceptable
                         if the  peak area remains within  +15% of the area obtained in the  initial
                         analysis of the standard.

7.     CALIBRATION

7.1    Assemble the HPLC and establish the operating  conditions in Table 2.
7.2    Retention time adjustment.
       7.2.1   Inject 20 \tL of the medium level calibration standard (Table  2).
       7.2.2   Locate the three peaks in this standard.

28

-------
                                                                                   Method 1654A
       7.2.3   Adjust the initial solvent mixture, the isocratic hold, the gradient, and the final
               isocratic hold until the retention times are within ± 1  minute of the retention times
               given in Table 2.
7.3    Minimum level: Analyze 20 juL of the low-level calibration standard (Table 2) and verify that
       the HPLC instrument meets the minimum level for phenanthrene in Table  1.
7.4    External standard calibration.
       7.4.1   Analyze 20 /iL of each calibration standard (Table 2) beginning with the lowest
               concentration and proceeding to the highest using to the procedure in Section 11.
       7.4.2   Record the areas for the phenanthrene peak and the height of the phenanthrene peak in
               the high-level standard.
       7.4.3   Compute the ratio of response to amount injected (calibration factor) at each
               concentration by dividing the area of the peak by the concentration of the standard
               injected.  Calculate the mean of the three values to produce an average calibration
               factor.
       7.4.4   Linearity: If the calibration factor  is constant over the three point calibration range
               (< 15% relative standard deviation), linearity through the origin can be assumed; if
               not, the system shall be recalibrated.
7.5    The average calibration factor  is verified on  each working 8-hour shift by the measurement of
       the medium-level calibration standard (Section 12.5).
7.6    Single-point calibration for diesel oil:  Inject the precision and recovery standard
       (Section 6.2.3) to produce a single calibration point for diesel oil.
       7.6.1   Integrate the area from the retention time of naphthalene (including the leading edge
               of the naphthalene peak) through the end of the indeno[l,2,3-of|pyrene peak or until
               the detector  signal returns to a stable baseline,  whichever comes later, as shown in
               Figure 1.
       7.6.2   Determine the calibration factor for diesel oil by dividing the integrated area
               (Section 7.6.1) by the diesel oil concentration (Section 6.2.1.1).

8.     QUALITY ASSURANCE/QUALITY CONTROL

8.1    Each laboratory that uses this method  is required to operate a formal quality assurance
       program (Reference 4). The minimum requirements of this program consist of an initial
       demonstration  of laboratory capability, an ongoing analysis of standards and blanks  as a test of
       continued performance, analyses of spiked samples to assess accuracy, and analysis of
       duplicates  to assess precision.  Laboratory performance is compared  to established
       performance criteria to determine if the results of analyses meet the performance
       characteristics  of the  method.   If the determination of PAH is to be made on extracts from
       Method  1662,  the quality control samples for initial precision and recovery (IPR), spiked
       samples, duplicate samples, and ongoing precision and recovery (OPR) samples from Method
       1662 shall be substituted for those in  the QC tests below,  and  the specifications in Table 1 for
       extracts from Method 1662 shall be met.
                                                                                             29

-------
Method 1654A
       8.1.1   The analyst shall make an initial demonstration of the ability to generate acceptable
               accuracy and precision with this method.  This ability is established as described in
               Section 8.2.
       8.1.2   The analyst is permitted to modify this method to improve separations or lower the
               costs of measurements, provided all performance requirements are met.  Each time a
               modification is made to the method, the analyst is required to achieve the minimum
               level (Section 7.3) and to repeat the procedure in Section 8.2 to demonstrate method
               performance.
       8.1.3   Analyses of spiked samples are required to demonstrate method accuracy when
               extracts from Method 1662 are analyzed. The procedure and QC criteria for spiking
               are described in Section 8.3.
       8.1.4   Analyses of duplicate samples  are required to demonstrate method precision when
               extracts from Method 1662 are analyzed.  The procedure and QC criteria for
               duplicates are described in Section 8.4.
       8.1.5   Analyses of blanks are required to demonstrate freedom from contamination.  The
               procedures and criteria for analysis of a blank are described  in Section 8.5.
       8.1.6   The laboratory shall, on an ongoing basis, demonstrate through calibration verification
               and analysis of the precision and recovery standard that the analysis system is in
               control.  These procedures are described in Section  12.5 and 12.6.
       8.1.7   The laboratory shall maintain records to define the quality of data that is generated.
               Development of accuracy statements is described in  Sections 8.3.2 and  12.6.4.
8.2   Initial precision and recovery (IPR):  The initial precision and recovery test is performed using
       the precision and recovery standard.  If extracts from Method 1662 are to be analyzed, the
       extracts from the initial precision and recovery tests in that method shall be used; otherwise,
       the laboratory shall generate acceptable precision and recovery by performing the following
       operations.
       8.2.1   Using diesel oil,  prepare four separate aliquots of the precision and  recovery standard
               (Section 6.2.3).  If extracts from Method 1662 are analyzed, the extracts from the
               initial precision and recovery test in that method shall be used.  Analyze these
               aliquots using the procedure in Section 11.
       8.2.2   Using results of the set of four analyses,  compute the average recovery (X) of PAH in
               mg/mL and the standard deviation of the recovery (s) in mg/mL for each aliquot by
               the external standard method (Sections 7.4 and 14.4).
       8.2.3   Compare s and X with the corresponding limits for initial precision  and recovery  in
               Table 1.  If s and X meet  the acceptance criteria, system performance is acceptable
               and analysis of oil samples may begin.  If, however, s exceeds the precision limit or
               X falls outside the range for accuracy, system performance  is unacceptable.  In  this
               event, review this method, correct the problem, and repeat the test.
8.3   Method accuracy:  If extracts from Method 1662 are to be analyzed,  the extract from the
       accuracy test in that method shall be used; otherwise, an accuracy test is unnecessary. The
       procedure for determining method  accuracy is given in Section 8.3 of Method 1662, and the
       specification for accuracy is given in Table 1 of this method.
30

-------
                                                                                   Method 1654A
       8.3.1   Compare the percent recovery of PAH with the corresponding QC acceptance criteria
               in Table 1.  If the results of the spike fail the acceptance criteria, and the recovery of
               the QC standard in the ongoing precision and recovery test (Section  12.6.3) is within
               the acceptance criteria in Table  1, an interference may be present. In this case, the
               result may not be reported for regulatory compliance purposes.  If, however, the
               results of both the spike and the ongoing precision and recovery test fail the
               acceptance criteria, the analytical system is judged to be out of control and the
               problem shall be identified and corrected, and the sample batch reanalyzed.
       8.3.2   As part of the QA program for the laboratory, method accuracy for  samples shall be
               assessed and records shall be  maintained.  After the analysis of five spiked samples
               in which the recovery passes  the test in Section 8.3, compute the average percent
               recovery (P) and the standard deviation of the percent recovery (sp).  Express the
               accuracy assessment  as a percent recovery interval from P  2sp to P +  2s,,. For
               example, if P  = 90% and sp  =  10% for five analyses of PAH in diesel oil, the
               accuracy interval is expressed as 70 to 110%.  Update the accuracy  assessment on a
               regular basis  (e.g., after each five to ten new accuracy measurements).
8.4    Duplicates: If extracts from Method 1662 are to be analyzed, the  extracts from  the duplicates
       test in that method shall be used.  The procedure for  preparing duplicates is given in Section
       8.4 of Method 1662, and the  specification for RPD is given in Table 1 of this method.  If
       extracts from Method  1662 are  not to  be analyzed, duplicates of the precision and recovery
       standard (Section 6.2.3) are analyzed,  and the specification for RPD is given  for PAH  in diesel
       oil in Table 1 of this method.
       8.4.1   Analyze each of the duplicates per the procedure in  Section 11 and compute the
               results per Section 14.
       8.4.2   Calculate the  relative percent  difference  (RPD) between the two results  per the
               following equation:


                                           Equation  1

                                    RPD=   ID'  " °2\  xlOO
                                           (D, + D2)/2

                where:
                D,  = Concentration  of diesel oil in the sample
                D^ = Concentration  of diesel oil in the second (duplicate) sample
       8.4.3   The relative percent difference for duplicates shall meet the acceptance criteria in
               Table 1.  If the criteria are not met, the analytical system is be judged to be out of
               control, and the problem must be immediately  identified and corrected and the sample
               set re-extracted and reanalyzed.
8.5    Blanks:  If extracts from Method  1662 are to be analyzed, the extracts from blanks in that
       method shall be analyzed in addition to the blanks in this method.
                                                                                              31

-------
Method 1654A
       8.5.1  Rinse the glassware used in preparation of the extracts in this method with acetonitrile
              and analyze a 20-^L aliquot of the rinsate using the procedure in Section 11 and
              compute the results per Section 14.
       8.5.2  If PAH is detected in a blank at greater than the method detection limit (MDL) in
              Table 1, analysis of samples is halted until the source of contamination is eliminated
              and a blank shows no evidence of contamination.
8.6   The specifications contained in this method can be met if the apparatus used is calibrated
       properly,  then maintained in a calibrated state.  The standards used for initial precision and
       recovery (IPR, Section 8.2) and ongoing precision and recovery (OPR, Section  12.6) should
       be identical, so that the most precise results will be obtained.  The HPLC instrument will
       provide the most reproducible results if dedicated to the settings and conditions required for
       the analyses given in this method.
8.7   Depending on specific program requirements, field  replicates and field spikes of diesel oil into
       samples may be required when Method 1662 and this method are used to assess the precision
       and  accuracy of the sampling and sample transportation techniques.

9.     SAMPLE COLLECTION, PRESERVATION, AND HANDLING

9.1    Oil samples are  collected in 20- to 40-mL vials with PTFE- or aluminum-foil-lined caps and
       stored in the dark at -20 to  -10°C.
9.2   If extracts from  Method 1662 are to be analyzed, the laboratory should be aware that sample
       and  extract holding times for this method have not yet been established. However, based on
       tests of wastewater for the analytes determined in this method, samples shall be extracted
       within 7 days of collection  and extracts shall be analyzed within 40 days of extraction.
9.3   As a precaution  against analyte and solvent loss or degradation, sample extracts are stored in
       glass bottles with PTFE-lined caps, in the dark, at -20 to -10°C.

10.   DILUTION OF  OIL AND EXTRACTS

10.1  Neat oil samples:  If oil is received in neat form, it should  be diluted to bring the
       concentration within the range of the instrument.  If the oil is No. 2 diesel oil, the appropriate
       concentration will be approximately 2000 /ig/mL.  Mineral oils and other oils containing a
       lesser PAH content will require less dilution.
10.2  Extracts from Method 1662:  If extracts of samples from Method 1662 are to be analyzed,
       these extracts (from Section 10.4.2 of that  method) are analyzed undiluted unless diesel oil is
       known or suspected to be present.  Extracts of QC  samples (IPR, OPR, matrix spikes, and
       duplicates) from Method  1662 are diluted by a factor of 10 to bring them within the range of
       the HPLC.
10.3  Dilution of neat oil expected to be diesel oil.
       10.3.1 Weigh  100 mg into a 10-mL volumetric flask and  dilute to the mark with acetonitrile
              to produce a concentration of 10 mg/mL.   Stopper and mix thoroughly.
       10.3.2 Using a calibrated  1.0-mL volumetric pipette, withdraw 1.0 mL of the  solution and
              place in a 10-mL volumetric flask. Then withdraw an additional 0.25 mL of the
32

-------
                                                                                 Method 1654A
               solution and place in the 10-mL volumetric flask (for a total of 1.25 mL).   Fill to the
               mark with acetonitrile to produce a concentration of 1.25 mg/mL (1250 jig/mL). This
               solution will be near, but not above, the limit of the calibration range and will match
               the concentration of the QC samples from Method 1662 (assuming 100% recovery).

7 /.  HIGH-PERFORMANCE LIQUID CHROMATOGRAPHY

11.1  Table 2 summarizes the recommended operating conditions for the HPLC.  Included in this
      table and in Table 1 are retention times and the minimum level that can be achieved under
      these conditions.  An example of the separation achieved for diesel oil by the multiple HPLC
      column system is shown in Figure 1.  Other HPLC columns, chromatographic conditions, or
      detectors may be used if the requirements for the minimum level (Sections 7.3)  and initial
      precision and recovery (Section 8.2) are met.
11.2 Calibrate the system as described in Section 7 or verify calibration as described  in Section 12.
11.3 Analysis of extracts.
      11.3.1  Inject 20 /xL of the sample extract, Method 1662 extract, or diluted QC extract into
               the HPLC using a high-pressure syringe or a constant-volume sample-injection loop.
               Record the volume injected to the nearest 0.1 /iL.
      11.3.2  Upon injection, begin the solvent program used in calibrating the column (Section
               7.2.3).  Record the signal from the time of injection until the detector returns to a
               stable baseline.  Return  the solvent to the initial conditions.
      11.3.3  Using the retention-time data determined during calibration, integrate the area from
               the retention time of naphthalene (including the leading edge of the naphthalene peak)
               through the end of the indeno[l,2,3-cd]pyrene peak or until the detector signal returns
               to a stable baseline, whichever comes later.
11.4 If the height of the response during the period recorded (Section  11.3.2) exceeds the height of
      the response for phenanthrene during calibration (Section 7.4.2), dilute the extract by
      successive factors of 10 with acetonitrile and reanalyze until the response is within the
      calibration range.

12.  HPL C SYSTEM AND LABORA TORY PERFORMANCE

12.1  At the beginning of each 8-hour  shift during which analyses are performed, HPLC calibration
      and system performance are verified.  For these tests, analysis of the medium-level calibration
      standard (Table 2) and of the diluted extract of the precision and recovery standard
      (Section 6.2.3) shall be used to verify all performance criteria.  Adjustment and/or
      recalibration (per Section 7) shall be performed  until all performance criteria are met.  Only
      after  all performance criteria are met may samples and blanks be analyzed.
12.2 Inject 20 /iL of the medium-level calibration standard (Table 2) into the HPLC instrument
      according to the procedure in Section 11.
12.3 Retention time:  The absolute retention times of the naphthalene, phenanthrene,  and
      indeno[l,2,3-o/]pyrene peaks shall be within +30 seconds of the respective retention times in
      the initial calibration (Section 7.2.3).
                                                                                            33

-------
Method 1654A
 12.4  HPLC resolution: Resolution is acceptable if the peak width at half-height of the phenanthrene
       peak is less than 30 seconds.
 12.5  Calibration verification:  Compute the concentration of phenanthrene based on the average
       calibration factor (Section 7.4.4).  The concentration shall be within the limits in Table 1.  If
       calibration is verified, system performance is acceptable and analysis of blanks and QC
       samples may begin.  If, however, the concentration falls outside of the calibration verification
       range, system performance is unacceptable.  In this case, correct the problem and repeat the
       test, or recalibrate (Section 7.4).
 12.6  Ongoing precision and recovery (OPR):  If the extract is from  Method 1662, the OPR standard
       from that method shall be used and the specification for the OPR from Method 1662 in
       Table 1  shall be met; if not,  a sample of diesel oil  shall be diluted per the procedure in
       Section  10 and shall be used  for the OPR test.
       12.6.1  Analyze the appropriate OPR standard.
       12.6.2 Compute the concentration of PAH in this standard per Section 14.
       12.6.3 Compare the concentration with the limits for ongoing precision and recovery in
               Table 1. If the concentration is in the range specified, the analytical processes are in
               control and  analysis  of blanks and  samples may proceed.  If, however, the
               concentration is not  in the specified range, these processes are not in control.  In this
               event, correct the problem, re-extract the sample batch if the OPR is from
               Method 1662, or redilute the oil sample (per Section 10.3), and repeat the ongoing
               precision and recovery test.
       12.6.4 Add results which pass the specification in Section 12.6.3 to initial and previous
               ongoing data. Update QC charts to form a graphic representation of continued
               laboratory performance.  Develop  a statement of laboratory data quality for each
               analyte by calculating the  average percent  recovery (R) and the standard deviation of
               percent recovery (sr).  Express  the accuracy as a recovery interval from R - 2sr to
               R +  2sr. For example, if R = 95% and sr = 5%, the accuracy is 85 to 105%.

 13.   QUALITATIVE IDENTIFICATION

 13.1  Qualitative determination is accomplished by comparison of data from  analysis of a sample or
       blank with data from  analysis of the calibration verification standard (Section 12.5).
 13.2  PAH is identified in  the sample by the presence of peaks and/or an elevated baseline (hump)
       between the retention times of the  naphthalene and indeno[l,2,3-ctf]pyrene peaks
       (Section 11.3.3), as shown in Figure 1.   The experience of the analyst shall weigh heavily in
       interpretation of the chromatogram.

 14.   QUANTITA TIVE DETERMINA TION

 14.1  Using the data system, compute the concentration of the PAH detected in the solution injected
       into the HPLC (in /ig/mL) using the calibration factor (Section 7.4).
 14.2  Concentration of PAH in oil:  If neat oil was analyzed, the concentration of PAH in the oil is
       determined  using the following equation:
34

-------
                                                                                 Method 1654A
                                         Equation 2
                                                C (itglmL)
                                   C  (mg/g)  = -!- -
                                    "            C. (mglmL)

               where:
               Co = Concentration of PAH in the oil sample
               Cp = Concentration of PAH measured (from Sections  11.4 and 14.1)
               C; = Concentration of oil in the solution injected into the HPLC
                    (from Sections  10.3.2, 11.4,  and 14.1)


14.3  Concentration of diesel oil in QC extracts from Method 1662: Calculate the concentration of
       diesel oil in QC extracts from Method 1662 by integrating the area per Section 7.6.1 and using
       the calibration from Section 7.6.2 of this method, taking into account the dilution of these
       extracts  (Section 10.2).
14.4  Concentration of PAH in oil from Method 1662: The PAH content of oil is complicated by
       the splitting and possible dilution of these extracts.
       14.4.1  Concentration in undiluted extracts: This concentration is determined by Equation 3:


                                         Equation 3
                                      V  xC      5 x  1 x C
                                   _   e    P    -             P_
                                  0  [1/SxWJ         Wr

                where:
                 Co = Concentration of PAH in the oil sample
                 V, = Amount of extract split for HPLC  analysis, in mL (1.0 mL)
                 Cp = Concentration of PAH measured
                 Wr = Weight of oil in the concentration  tube in Method 1662
                      (Section  11.5.5 of Method  1662)
                1/5 = Fraction of this  weight used for the PAH determination
       14.4.2 Concentration in diluted extracts:  If the extract was diluted by a factor of 10
               (Section 10.3 or 11.4), the concentration determined in Section 14.4.1 is multiplied
               by 10.
14.5  If the concentration is to be expressed as weight percent, C0 is multiplied by 0.1.
14.6  Report  results to three significant figures without correction for recovery.

15.   METHOD PERFORMANCE
This method was validated in a single laboratory (Reference 6) using samples of hot-rolled drilling
mud (Reference 7).

-------
Method 1654A
                                    References

1.    "Carcinogens—Working With Carcinogens." Department of Health, Education, and Welfare,
      Public Health Service, Centers for Disease Control [available through National Technical
      Information System, 5285 Port Royal Road, Springfield, VA 22161, document no. PB-
      277256]: August 1977.

2.    "OSHA Safety and Health Standards, General Industry [29 CFR 1910], Revised."
      Occupational Safety and Health Administration, OSHA 2206. Washington, DC: January 1976.

3.    "Safety  in Academic Chemistry Laboratories (3rd Edition)."  American Chemical Society
      Publication, Committee on Chemical Safety. Washington, DC:  1979.

4.    "Handbook of Analytical Quality Control in Water and Wastewater Laboratories." USEPA,
      EMSL-Ci, EPA-600/4-79-019. Cincinnati, OH: March 1979.

5.    "Standard Practice for Sampling Water," ASTM Annual Book of Standards, Part 31, D3370-
      76, ASTM.  Philadelphia, PA: 1980.

6.    "Determination of Polynuclear Aromatic Hydrocarbons and Diesel by Modified EPA Method
      8310."  Prepared for the American Petroleum Institute c/o Shell Development Co, Westhollow
      Research Center, 3333 Highway 6 South, Houston, TX 77082 by Analytical Technologies
      Inc., 225 Commerce Drive, Fort Collins, CO 80524: March 29 1991, April 12, 1991, and
      August  18, 1992.

7.    "Results of the API Study of Extraction and Analysis Procedures for the Determination of
      Diesel Oil in Drilling Muds (Final Report)." American Petroleum Institute, Offshore Effluent
      Guidelines Steering Committee, Technology Work Group, Prepared by J.C. Raia, Shell
      Development Co.  Houston, TX:  April 18,  1991.
36

-------
                                                                            Method 1654A
Table 1.  Performance Data and Method Acceptance Criteria for PAH
                                           PAH in
                                          Diesel Oil1
                                             100
                                             7.6
                          Diesel Oil in Mud
                              Extract2
Phenanthrene
     0.1
                                                                           0.39-0.61
          Criterion             Units
Minimum level3                  /yg/mL
Method Detection Limit4          //g/mg
Initial prec and recov
  Precision (std dev)
      PAH in diesel oil6          mg/mL         120
      Diesel in mud extract6      mg/mL          --             0.55
  Recovery
      PAH in diesel oil5          mg/mL     1090-1340
      Diesel in mud extract6      mg/mL          --          0.84-1.95
Calibration verification7           fjg/mL
Ongoing prec and recov
  PAH in diesel oil6             mg/mL     1010-1450
  Diesel  in mud extract          mg/mL          --          0.76-2.15
Matrix spike recovery6              pet           --          0.43-2.39
Duplicates                       RPD         9.5             44
Notes:
1 CAS Registry number 68534-30-5; No.  2 diesel oil used for these tests
2 From Method 1662
3 This is a minimum level at which the analytical system shall give recognizable signals and
  acceptable calibration points.
4 40 CFR Part 136, Appendix B; MDL is measured as PAH in oil
5 Test concentration of diesel oil =  1250 //g/mL
6 Test concentration in diluted extract =  1.25 mg/mL
7 Test concentration  = 0.50//g/mL
Table 2. HPLC Calibration Data
 Analyte

 Naphthalene

 Phenanthrene
 IndenoM ,2,3-ceflpyrene

 Diesel oil
Retention Time*
    (minutes)

       7.2

      10.0

      18.9

    7.0-20.0
                                                         Calibration solution
                                                        concentration (ug/mL)
Low

0.1


Medium
5
0.5
0.5
1250
High

2.0


   Column system:  Two C18 columns (150 mm long by 4.6 mm i.d., 300 angstroms) connected in
   series, preceded by one C1B guard column  (30 mm long by 4.6 mm i.d., 300 angstroms).
   Column temperature 30°C; solvent flow rate 1.5 mL/min; linear gradient from 50% water/
   50% acetonitrile  at injection to 100% acetonitrile in 10 minutes, hold at 100% acetonitrile for
   15 minutes.
                                                                                      37

-------
Method 1654A
                                    00
                                    O>

                                    O>
                                     I
                          oo

                          r^
                            _

                            CO
                            Q.
                            CO
                          Three-Component Standard
                                       03
                                       CO

                                       CD
                                                    0)
                                                    n

                                                    £
                                                    >,
                                                    Q.

                                                    f

                                                    CO
                                            No. 2 Diesel Oil
0.00
0.50
1.00              1.50


   x 10^ minutes
2.00
Figure  1.     Liquid Chromatography of the Three-Component Standard and of No. 2

             Diesel Oil
38

-------
         Method 1662
Total Extractable Material in
              Drilling Mud
    by SDS Extraction and
              Gravimetry

-------
                                   Method 1662
                   Total Extractable Material in Drilling Mud
                       by SDS Extraction and Gravimetry
1.    SCOPE AND A PPLICA TION

1.1   This method is designed to determine the oil content of drilling mud by Soxhlet/Dean-Stark
      (SDS) extraction and gravimetric measurement.  However, this is a method-defined
      measurement that does not discriminate oil from other materials capable of being extracted
      from the mud.  EPA Methods 1651, 1654, and 1663 can be used to aid in determining the
      presence and identity of oil.
1.2   This method is for use in the Environmental Protection Agency's (EPA's) survey and
      monitoring programs under the Federal Water Pollution Control Act.
1.3   The detection limit of this method is usually dependent on the level of background materials in
      the drilling mud rather than instrumental limitations. The level in Table  1 typifies the
      minimum level  that  can be detected with no interferences present.
1.4   Any modification of this method beyond those expressly permitted shall be considered a major
      modification subject to application and approval of alternative test procedures under
      40 CFR 136.4 and  136.5.
1.5   Each  analyst that uses this method must demonstrate the ability to generate acceptable results
      using the procedure in Section 8.2.

2.    SUMMARY OF METHOD

2.1   Approximately 25 g of drilling mud is extracted with toluene in an SDS extractor
      (Reference 1).  The extract is evaporated to dryness using a rotary evaporator and nitrogen
      blowdown apparatus. The weight of oil is determined using an analytical balance.
2.2   Quality is assured through reproducible calibration and testing of the extraction and
      gravimetric systems.

3.    INTERFERENCES

3.1   Solvents, reagents, glassware, and other sample processing hardware may lead to discrete
      artifacts and elevated measurements causing misinterpretation of results.
3.2   All materials used in the analysis shall be demonstrated to be free from interferences by
      running method blanks initially and with each sample batch (samples started through the
      extraction process at the same time, to a maximum of ten).  Specific selection of reagents and
      purification of solvents by distillation in all-glass systems may be required.
3.3   Glassware and, where possible, reagents are cleaned by solvent rinse and/or baking at 200°C
      for a minimum  of 1  hour.
                                                                                         41

-------
 Method 1662
 4.    SAFETY

 4.1   The toxicity or carcinogenicity of each reagent used in this method has not been precisely
       defined; however, each chemical should be treated as a potential health hazard.  Exposure to
       these chemicals must be reduced to the lowest possible level.
 4.2   The laboratory is responsible for maintaining a current awareness file of OSHA regulations
       regarding the safe handling of the chemicals specified in this method. A reference file of
       material safety data sheets (MSDSs) should be made available to all personnel involved in the
       chemical analysis.  Additional information on laboratory safety can be found in References 2
       through 4.
 4.3   Methylene chloride has been classified as a known health hazard.  All steps in this method
       which involve exposure to this compound shall be performed in an OSHA-approved fume
       hood.

 5.    APPARA TUS AND MA TERIALS

       NOTE:  Brand names, suppliers, and part numbers are for illustrative purposes only.
       No endorsement is implied. Equivalent performance may be achieved using apparatus
       and materials other than those specified here, but demonstration of equivalent
       performance meeting the requirements of this method is the responsibility of the
       laboratory.

 5.1   Sampling equipment for discrete sampling.
       5.1.1   Sample bottle: Wide-mouth amber glass or opaque cosmetic jars,  100-mL minimum,
               with screw-cap.  If amber bottles are not available, samples  shall be protected from
               light.
       5.1.2   Bottle caps:  Threaded to fit sample bottles.  Caps shall be lined with PTFE or
               aluminum.
       5.1.3   Cleaning.
               5.1.3.1   Bottles: Detergent water wash, tap water rinse, cap with aluminum foil,
                         and bake at 110 to 200°C for a minimum of 1 hour prior to use.
               5.1.3.2   Liners:  Detergent wash, tap water and solvent rinse, and bake at 110 to
                         200°C for a minimum of 1 hour prior to use.
       5.1.4   Bottles and liners must be lot-certified to be free of artifacts by running blanks
               according to this method.  If blanks from bottles and/or liners without cleaning or
               with fewer cleaning steps than required above show no detectable materials (per
               Section 8.5), the bottle and liner cleaning steps that do not eliminate these artifacts
               may be omitted.
 5.2   Equipment for glassware cleaning.
       5.2.1   Laboratory sink with overhead fume hood.
       5.2.2   Oven: Capable of reaching 200°C within 2 hours and holding 200°C within ± 10°C.
42

-------
                                                                                   Method 1662
5.3   Equipment for sample preparation.
       5.3.1   Laboratory fume hood.
       5.3.2   Balances.
               5.3.2.1   Analytical: Capable of weighing 1.0 mg.
               5.3.2.2   Top loading:  Capable of weighing 100 mg.
       5.3.3   Beaker:  400- to 500-mL.
       5.3.4   Spatula: Stainless steel.
       5.3.5   Desiccator:  Cabinet- or jar-type, capable of keeping the Kuderna-Danish concentrator
               tubes (Section 5.5.1.1) dry during cooling.
5.4   Soxhlet/Dean-Stark (SDS) extractor.
       5.4.1   Soxhlet: 50 mm i.d., 200-mL capacity with 500-mL flask (Cal-Glass LG-6900, or
               equivalent, except substitute 500-mL round-bottom flask for 300-mL flat-bottom
               flask).
       5.4.2   Thimble:   123 mm long by 43 mm i.d. to fit Soxhlet (Cal-Glass LG-6901-122, or
               equivalent).
       5.4.3   Moisture trap:  Dean-Stark or Barret with PTFE stopcock to fit Soxhlet (Figure 1).
       5.4.4   Heating mantle: Hemispherical, to fit 500-mL round-bottom flask (Cal-Glass
               LG-8801-112, or equivalent).
       5.4.5   Variable transformer:  Powerstat (Cal-Glass LG-8965-100, or equivalent).
5.5   Concentration apparatus.
       5.5.1   Rotary evaporator or other concentration device capable of evaporating toluene: Any
               concentration technique may be used provided that the  requirements of Section 8.2
               and the method detection limit in Table 1 are met.
       5.5.2   Rotary evaporator with vacuum pump.
               5.5.1.1   Rotary evaporator:  Operated at approximately 60 rpm with built-in water
                         bath operated at approximately 90°C and condenser  with tap water
                         approximately 45°C (Buchi Model Re-121,  or equivalent).
               5.5.1.2   Vacuum pump:  CAST Model 1HAP-25-M100X (or equivalent).
       5.5.2   Nitrogen blowdown apparatus:  Equipped with water bath controlled at 40 to 50°C
               (N-Evap, Organomation Associates, Inc., or  equivalent), installed in a fume hood.
               5.5.2.1   Concentrator tube:   10-to 15-mL, graduated (Kontes K-570050-1025, or
                         equivalent)  with calibration verified.
       5.5.3   Pipettes.
               5.5.3.1   Disposable, Pasteur, 150 mm long by 5 mm i.d. (Fisher Scientific
                         13-678-6A, or equivalent).
               5.5.3.2   Disposable, serological, 10-mL (6 mm  i.d.).
5.6    Calculator or computer:  Capable of calculating and maintaining statistics on initial
       (Section 8.2) and ongoing (Section 11.3.4) performance.
                                                                                           43

-------
Method 1662
6.     REAGENTS

6.1    Reference matrix:  Blank drilling mud, playground sand, or similar material in which the
       compounds of interest and interfering compounds are not detected by this method.  May be
       prepared  by pre-extraction with methylene chloride and drying at  110 to 200°C for a minimum
       of 4 hours.
6.2    Solvent:  Toluene and methylene chloride, distilled in glass (Burdick and Jackson, or
       equivalent).
6.3    White quartz sand, 60/70 mesh:  For Soxhlet/Dean-Stark extraction, (Aldrich Chemical Co,
       Cat No. 27,437-9, or equivalent).
6.4    Standard  for diesel oil: Ideally, the oil standard used in this method should be from the oil
       used on the drilling rig from which the mud sample is to be taken. If this oil is not available,
       No. 2 diesel oil may be substituted.  When not being used, the standard is stored in the dark at
       -20 to -10°C in a screw-capped vial with PTFE-lined lid.  A mark is placed on the vial at the
       level of the solution so that solvent loss by evaporation can be detected. The vial is brought to
       room temperature prior to use and solvent is added (if required).
6.5    Stock solution.
       6.5.1   Diesel oil  in  toluene: Weigh 6.25 g of diesel oil to three significant figures in a 100-
               mL ground-glass stoppered volumetric flask and fill to the mark with toluene.  After
               the oil is completely dissolved, transfer the solution to a clean  150-mL bottle with
               PTFE-lined cap.
       6.5.2   The stock solution should be checked  for signs of degradation  prior to the preparation
               of the precision and recovery standard.
6.6    Precision and recovery standard:  The stock solution is spiked into the reference matrix
       (Section 6.1) for the determination of initial  precision and recovery (IPR; Section 8.2) and
       ongoing precision and recovery (OPR; Section  11.2).  When 1 mL of this  solution is spiked
       into a 25-g reference  matrix sample, a concentration of 0.25% (2.5 g/kg) will be produced.

7.     CALIBRA TION

7.1    Verify calibration of the balance at 10 mg and 100 mg using class "S" weights.
7.2    Calibration shall  be within + 10% (1 mg) at 10 mg and  +2% (2 mg) at 100 mg.  If not within
       these limits, recalibrate the balance and  repeat the test.

8.     QUALITY ASSURANCE/QUALITY CONTROL

8.1    Each laboratory that uses  this method is required to operate a formal quality assurance
       program (Reference 5).  The minimum requirements of this program consist of an initial
       demonstration of laboratory capability, an ongoing analysis of standards and blanks as a test  of
       continued performance, analyses of spiked samples to assess accuracy,  and analysis of
       duplicates to assess precision. Laboratory performance  is compared to  established
       performance criteria to determine if the results of analyses  meet the performance
       characteristics of the method.
44

-------
                                                                                    Method 1662
       8.1.1   The analyst shall make an initial demonstration of the ability to generate acceptable
               accuracy and precision with this method.  This ability is established as described in
               Section 8.2.
       8.1.2   The analyst is permitted to modify this method to improve separations or lower the
               costs of measurements, provided all performance requirements are met.  Such
               modifications may use alternative extraction or concentration techniques or alternative
               HPLC columns. Each time a modification is made to the method, the analyst is
               required to repeat the procedure in Section 8.2 to demonstrate method performance.
       8.1.3   Analyses of spiked samples are required to demonstrate method accuracy.  The
               procedure and QC criteria for spiking are described in Section 8.3.
       8.1.4   Analyses of duplicate samples are required to demonstrate method precision. The
               procedure and QC criteria for duplicates are described in Section 8.4.
       8.1.5   Analyses of blanks are required to demonstrate freedom  from contamination.  The
               procedures and criteria for analysis of a blank are described in Section 8.5.
       8.1.6   The laboratory  shall, on an ongoing basis, demonstrate through calibration verification
               and analysis of the OPR sample that the analysis system  is in  control.  These
               procedures are described in Section 11.
       8.1.7   The laboratory  shall maintain records to define the quality of data that is generated.
               Development of accuracy statements is described in Sections 8.3.4 and  11.2.4.
8.2    Initial precision and accuracy (IPR):  To establish the ability to generate acceptable precision
       and accuracy, the analyst shall perform the following operations.
       8.2.1   Extract and evaporate four samples of the precision and recovery standard
               (Section 6.6)  according to the procedure beginning in Section 10.
       8.2.2   Using results of the set of four  analyses, compute the average recovery (X)  in g/kg
               and the standard deviation of the recovery (s) in g/kg for each sample.
       8.2.3   For each compound, compare s and X with the corresponding limits for initial
               precision and accuracy in Table 1. If s  and X meet the acceptance criteria,  system
               performance is  acceptable and  analysis of samples may begin.  If, however, s exceeds
               the precision  limit or X falls outside the range for accuracy, system performance is
               unacceptable.  In this  event, review this method, correct the problem, and repeat the
               test.
8.3    Method  accuracy:  The laboratory shall  spike a minimum of 10%  (one sample in each set of
       ten samples) of all drilling mud samples. This sample shall be spiked  with the diesel oil that
       was added to the drilling fluid.  If a reference standard of diesel oil that was added to the
       drilling fluid is not available,  No. 2 diesel oil shall be used for this spike.  If doubt of the
       concentration of diesel oil  in any of the remaining 90% of the samples exists, that sample shall
       be spiked to confirm the diesel oil concentration.
       8.3.1   The concentration of the spike in the sample shall be determined  as follows.
               8.3.1.1   If, as in compliance monitoring, the  concentration of the oil in the sample
                         is  being checked against a regulatory concentration limit, the spike shall be
                         at that limit or  at 1  to 5 times higher than the background concentration
                         determined in Section 8.3.2,  whichever concentration is higher.

                                                                                             45

-------
Method 1662
               8.3.1.2   If the concentration of the oil in a sample is not being checked against a
                          limit, the spike shall be at the concentration of the precision and recovery
                          standard (Section 6.6) or  at 1 to 5 times higher than the background
                          concentration, whichever  concentration is higher.
       8.3.2  Analyze one sample aliquot to determine the background concentration (B) of oil.  If
               necessary,  prepare a standard solution appropriate to produce a level in the sample at
               the regulatory concentration limit or  at 1 to 5 times the background  concentration (per
               Section 8.3.1).  Spike a second sample aliquot with the standard solution and analyze
               it to determine the concentration after spiking (A) of each analyte.  Calculate the
               percent recovery (P) of the oil using  Equation 1:


                                          Equation  1
                                         p  =  100 (A-B)
                                                  f
                            where'.
                            A  =  Concentration of analyte after spiking
                            B  =  Background concentration of oil
                            T  =  True value  of the  spike
       8.3.3  Compare the percent recovery for total oil with the corresponding QC acceptance
               criteria in Table 1.  If the results of the spike fail the acceptance criteria, and the
               recovery of the QC standard in the ongoing precision and recovery test (Sections
               10.1.3 and 11.2.4) is within the acceptance criteria in Table 1, an interference may be
               present.  (See Section 3 for identification of interferences).  In this case, the result
               may not be reported for regulatory compliance purposes. If, however, the results of
               both the spike and  the ongoing precision and recovery test fail  the acceptance criteria,
               me analytical system is judged to be out of control, and  the problem shall be
               identified and corrected  and the sample batch  reanalyzed.
       8.3.4  As part of the QA  program for the laboratory, method accuracy for samples shall be
               assessed and records shall be maintained.  After the analysis of five spiked samples
               in  which the recovery passes the test in Section 8.3, compute the average percent
               recovery (P) and the standard deviation of the percent recovery (sp).  Express the
               accuracy assessment as a percent recovery interval from P  2sp to P + 2sp.  For
               example, if P = 90% and sp = 10% for five analyses of diesel oil, the accuracy
               interval  is expressed as 70 to 110%.  Update the accuracy assessment  on a regular
               basis (e.g., after each five to ten new accuracy measurements).
8.4   The laboratory shall analyze duplicate samples for each drilling-mud type at a minimum of
       10% (one sample for each ten sample set).  A duplicate sample shall consist of a well-mixed,
       representative aliquot of the sample.
       8.4.1   Analyze one sample in the set in duplicate per the procedure beginning in Section 10.
       8.4.2   Compute the relative percent difference (RPD) between  the two results per the
               following equation:
46

-------
                                                                                  Method 1662
                                         Equation 2

                                   RPD =   I  '  	d_ x 100
                                         (D, + Z>2)/2

                where:
                £>,  =  Concentration of diesel in the sample
                D2  =  Concentration of diesel oil in the second (duplicate) sample


       8.4.3  The relative percent difference for duplicates shall meet the acceptance criteria in
              Table 1.  If the criteria are not met, the analytical system is be judged to be out of
              control, and the problem must be immediately identified and corrected and the sample
              set reanalyzed.
8.5    Blanks: Reference matrix blanks (Section 6.1) are analyzed to demonstrate freedom from
       contamination.
       8.5.1  Extract and concentrate a 25-g aliquot of the reference matrix initially and with each
              sample batch (samples started through the analysis at the same time, to a maximum of
              ten samples).
       8.5.2  If greater than 0.2 g/kg of material is detected in a blank, analysis of samples is
              halted until the source of contamination is eliminated and a blank shows no evidence
              of contamination.
8.6    The specifications contained in this method can be met if the apparatus used is calibrated
       properly,  then maintained  in a calibrated state. The standards used for initial precision and
       recovery (IPR, Section 8.2) and ongoing precision and recovery (OPR, Section 11.2) should
       be identical, so that the most precise results will be obtained.
8.7    Depending on specific program requirements,  field replicates and field spikes of diesel oil into
       samples may be required to assess the precision and accuracy of the sampling and sample
       transporting techniques.

9.     SAMPLE COLLECTION,  PRESERVATION,  AND HANDLING

9.1    Collect drilling mud samples in wide-mouth glass containers following conventional sampling
       practices (Reference 6).
9.2    Samples must be representative of the entire bulk drilling mud.  In some instances, composite
       samples may be  required.
9.3    Maintain samples in the dark at 0 to 4°C  from the time of collection until analysis.
9.4    Analyze samples within 28 days of collection.

10.    SAMPLE EXTRACTION  AND CONCENTRATION

10.1   Preparation of sample and  QC aliquots.
       10.1.1 Transfer approximately 25 g of a well-homogenized and representative portion of the
              drilling mud to a tared 400- to 500-mL beaker.  Determine and record the weight.
                                                                                           47

-------
Method 1662
       10.1.2 Add approximately 50 g of quartz sand (Section 6.3) to the beaker and mix the
               drilling mud and sand thoroughly.
       10.1.3 QC standard and blank:  Used for tests of initial (Section 8.2) and ongoing
               (Section 11.2) precision and accuracy.  For each of the four initial precision and
               recovery (IPR) standards, the ongoing precision and recovery (OPR) standard, and the
               blank (Section 8.5), prepare aliquots as follows.
               10.1.3.1  Place approximately 25 g of the reference matrix (Section 6.1) in a clean
                         400- to 500-mL beaker.
               10.1.3.2  Spike 1 mL of the precision and recovery standard (Section 6.6) into the
                         IPR or OPR standard.  Do not spike the blank.
               10.1.3.3  Add  approximately 50 g of quartz sand (Section 6.3) to the beaker and mix
                         thoroughly.
10.2  Soxhlet/Dean-Stark extraction.
       10.2.1  Pre-extraction:   Used to clean the SDS extractor.  Pre-extraction may be  eliminated if
               extractable material is not found in blanks.
               10.2.1.1  Charge a clean  extraction thimble with 50 g of quartz sand (Section 6.3).
                         Do not disturb the silica layer throughout the extraction process.
               10.2.1.2  Place the thimble in a clean extractor.  Place 30 to 40 mL of toluene in the
                         receiver and 200 to 250 mL in the flask.
               10.2.1.3  Begin the extraction by heating the flask until the toluene is boiling.  When
                         properly adjusted, one to  two  drops  of toluene per second will fall from
                         the condenser tip into the receiver.   Pre-extract the apparatus for a
                         minimum of 4 hours.
               10.2.1.4  After pre-extraction,  cool and disassemble the apparatus.   Rinse with
                         methylene chloride and allow  to air-dry in a hood.
       10.2.2 SDS extraction.
               10.2.2.1  Load  the sample (from Section 10.1.2) and QC aliquot(s) and  blank (from
                         Section 10.1.3.3) into pre-cleaned thimbles.
               10.2.2.2  Reassemble the apparatus and add a  fresh charge of toluene to the
                         receivers and reflux flasks.  Rinse the beakers into their respective
                         thimbles using 10 to 20 mL of toluene.
               10.2.2.3  Apply power to the heating mantle to begin refluxing.   Adjust the reflux
                         rate to match the rate of percolation  through the sand bed until sufficient
                         water has been  removed so that the flow of toluene is no longer restricted.
               10.2.2.4  Drain the water from the receiver at 1  to 2 hours and 8 to 9 hours, or
                         sooner if the receiver fills with water.  Record the total volume of water
                         collected.  Reflux the sample  for a total of 16 to 24 hours.  Cool to room
                         temperature.
               10.2.2.5  Estimate and record the volume of extract (to the nearest  100  mL).
48

-------
                                                                                   Method 1662
10.3 Concentration.
      10.3.1 Add one or two clean boiling chips to the evaporative flask and attach it to the rotary
              evaporator.
              10.3.1.1  Place the round-bottom flask in the hot-water bath (approximately 90°C)
                         so that approximately one-half of the flask is immersed in hot water.
              10.3.1.2  Start the flow of cooling water and turn on the vacuum pump.  Rotate the
                         flask slowly at first to control the rate of evaporation of the toluene.
              10.3.1.3  When the apparent volume reaches a few  milliliters, remove the flask from
                         the hot water bath and allow it to cool for at least 10 minutes.  To
                         minimize the loss of the more volatile components of oil, do not take the
                         sample to dryness in the rotary evaporator.
              10.3.1.4  Turn off the pump and cooling water.
              10.3.1.5  Disassemble the apparatus.
              10.3.1.6  Using a squeeze bottle or pipette, rinse the inside surface of the round-
                         bottom flask with a small  portion of acetonitrile.  Transfer the solution to a
                         calibrated Kuderna-Danish concentrator tube. Repeat the rinsing and
                         transfer two more times to quantitatively transfer the solution to the
                         concentrator tube.
10.4 Extract for other analyses:  If a portion of the extract is to be  retained for HPLC analysis
      using Method 1654A or GC/FID analysis using Method 1663, the extract is split as follows.
      10.4.1 Adjust the  extract  volume to 5.0 mL with acetonitrile.
      10.4.2 Remove 1.00 mL with a volumetric pipette and place in a 2- to 3-mL amber vial with
              PTFE-lined screw-cap.  Mark a line on the vial at the level of the solution. Seal and
              store in the dark at -20 to -10°C.
      10.4.3 Evaporate the remaining 4 mL of extract per the steps below.
10.5 Evaporation to dryness.
      10.5.1 Place the receiver  in the water in the nitrogen blowdown apparatus.  Adjust the water
              temperature to 40 to 50°C.
      10.5.2 Adjust the  height of the blowdown needle to approximately 1 cm above the surface of
              the solution.
      10.5.3 Adjust the  nitrogen flow rate so that it is sufficient  to create a depression in the
              surface  of the solution but not so great that the solution spatters.
      10.5.4 Evaporate the solvent until the volume is constant,  but no longer than 30 minutes.
              Wipe the outside surface of the concentrator tube dry and cool the tube in the
              desiccator.
      10.5.5 Weigh the  receiver.  If doubt exists that constant weight has been achieved, return the
              receiver to the blowdown  apparatus and evaporate solvent for 15 minutes more.  Cool
              the receiver in the desiccator and weigh the receiver.   Constant weight is achieved
              when the readings differ by less than 5% or 5 mg,  whichever is greater.
                                                                                            49

-------
Method 1662
 71.   SYSTEM AND LABORA TORY PERFORMANCE

 11.1   Calibration verification:  Verify calibration of the balance per Section 7 before and after each
       set of 12 or fewer measurements. (The 12 measurements will normally be ten samples, plus
       one ongoing precision and recovery  standard,  plus one blank.) If calibration is not verified
       after  the measurements, recalibrate the balance and reweigh the batch.
 11.2  Ongoing precision and recovery.
       11.2.1 Weigh the evaporated precision and recovery standard extracted and concentrated with
              each batch of samples.
       11.2.2 Calculate the concentration  of oil in this standard.
       11.2.3 Compare the concentration  with the limits for ongoing precision and recovery in
              Table 1.  If the concentration is in the range specified, the extraction and evaporation
              processes are in control and analysis of blanks and samples may proceed.  If,
              however, the concentration  is not in the specified range, these processes are not in
              control.  In  this  event, correct the problem, re-extract the sample batch, and repeat
              the ongoing precision and recovery test.
       11.2.4 Add results  that pass the specification in Section 11.2.3 to initial and previous ongoing
              data.  Update QC charts to  form a graphic representation of continued laboratory
              performance.  Develop a statement of laboratory data quality for each analyte by
              calculating the average percent recovery (R) and the standard deviation of percent
              recovery (sr).  Express the accuracy as a recovery interval from R   2sr to R  + 2sr.
              For example, if R = 95% and sr = 5%, the accuracy is 85 to 105%.

 12.   QUANTITATIVE DETERMINATION

 12.1   Determination of dry weight of sample and of percent solids.
       12.1.1 Using the sample weight (Section 10.1.1)  and the weight (volume) of water in the
              moisture trap (Section 10.2.4.4), calculate the dry weight of solids in the sample as
              follows:


                                         Equation 3
                                        w=w-w
                                         d     s      w
                              where:
                              Wd = Dry  weight of solids,  in  grams
                              Ws =  Weight of sample,  in  grams
                              Ww =  Weight of water, in grams
50

-------
                                                                                 Method 1662
       12.1.2 Calculate the percent solids as follows:
                                         Equation 4
                                                     W.
                                     % solids = 100  -J.
                                                     ^
                             where:
                             Wd = Dry weight of solids, in grams
                             Ws = Weight of sample,  in grams
12.2 Determination of extractable material concentration.
      12.2.1 Calculate the concentration of extractable material in the total (wet) sample using the
              following equation:
                                         Equation 5
                                                         W
                                 Concentration (%) = 0.1 —
                                                         Ws
                     where:
                     Wr = Weight of extractable material in  receiver, in mg
                     Ws = Weight of sample, in grams

       12.2.2 If 1 mL of the 5-mL extract was removed for HPLC or GC analysis, multiply the
              result by 1.25 to compensate for this loss.
12.3  Report results to two significant figures without correction for recovery.

13.   METHOD PERFORMANCE
This method was validated in a single laboratory (Reference 7) using samples of hot-rolled drilling
mud (Reference 8).
                                                                                          51

-------
Method 1662
                                    References

1.     Lamparski, L.L. and T.J. Nestrick. "Novel Extraction Device for the Determination of
       Chlorinated Dibenzo-p-dioxins (PCDDs) and Dibenzofurans (PCDFs) in Matrices Containing
       Water."  Chemosphere, 19: 27-31, 1989.

2.     "Carcinogens—Working With Carcinogens."  Department of Health, Education, and  Welfare,
       Public Health Service, Centers for  Disease Control [available through National Technical
       Information System, 5285 Port Royal Road, Springfield, VA 22161, document no. PB-
       277256]: August 1977.

3.     "OSHA Safety and Health Standards, General Industry [29 CFR 1910], Revised."
       Occupational Safety and Health Administration, OSHA 2206.  Washington, DC: January 1976.

4.     "Safety in Academic Chemistry Laboratories (3rd Edition)."  American Chemical Society
       Publication. Committee on Chemical Safety.  Washington, DC: 1979.

5.     "Handbook of Analytical Quality Control in Water and Wastewater Laboratories." USEPA,
       EMSL-Ci, EPA-600/4-79-019.  Cincinnati, OH: March 1979.

6.     "Standard Practice for Sampling Water," ASTM Annual Book of Standards, Part 31,  D3370-
       76, ASTM, Philadelphia, PA:  1980.

7      "Results of the API Study of Extraction and Analysis Procedures for the Determination of
       Diesel Oil in Drilling Muds (Final  Report)."  American Petroleum Institute, Offshore Effluent
       Guidelines Steering Committee, Technology Work Group, Prepared by J.C. Raia, Shell
       Development  Co. Houston, TX:  April  18, 1991.

8.     "Development of Specifications for Method 1662."  Analytical Technologies, Inc., Work
       Order 92-06-025, prepared for the  American Petroleum Institute.  1220 L St NW, Washington,
       DC: August 18,  1992.
52

-------
                                                                              Method 1662
Table 1.  Method Acceptance Criteria for Diesel Oil1  in Drilling Mud
                                                                   Diesel OH
       Acceptance Criterion
Method Detection Limit (matrix)3
Initial precision and recovery4
   Precision [standard deviation]
   Recovery [mean; X]
Ongoing precision and recovery4
Matrix spike recovery
Precision of duplicates

Notes:
Section

 8.2.3
 11.2.3
 8.3.3
 8.4.3
Units2
 g/kg

 g/kg
 g/kg
 g/kg
percent
 RPD
 Amount
    1.1

   0.85
1.18 -3.73
1.08 -3.83
 35   159
    34
 1  CAS Registry number 68534-30-5
 2  To convert to weight percent, multiply the amount by 0.1
 3  40 CFR Part 136, Appendix B; measured in API Mud number 101-1
 4  Text concentration:  2.5 g/kg (0.25 percent) diesel oil in mud
 5  Precision of duplicate analyses must be <34%.
                                                                                       53

-------
Method 1662
                                                                                 AS2-010-002
Figure 1.  Soxhlet/Dean-Stark Extractor




54

-------
         Method 1663
Differentiation of Diesel and
       Crude Oil by GC/FID

-------
                                   Method  1663
              Differentiation of Diesel and Crude Oil by GC/FID
1.    SCOPE AND APPLICA TION

1.1   This method differentiates between diesel oil and crude oil in drilling muds and other sources
      by comparing the ratio of n-alkanes in the Cg-C^ range as determined by gas chromatography
      (GC) with a flame-ionization detector (FID).
1.2   This method is for use in the Environmental Protection Agency's survey programs and may be
      used for compliance monitoring as part of the "Effluent Limitations Guidelines and New
      Source Performance Standards for the Offshore Subcategory of the Oil and Gas Extraction
      Point Source Category" [50 FR 34592].
1.3   For oil in drilling muds, this method is designed to be used in  conjunction with the extraction
      procedure in EPA Method 1662.
1.4   This method cannot differentiate between mineral oil and diesel/crude oil.  EPA
      Method 1654A can be used to determine that the oil is not mineral oil by measurement of the
      polynuclear aromatic (PAH) content.
1.5   When used in conjunction with EPA Method 1662, the estimated detection limit for diesel or
      crude oil in drilling mud is  100 mg/kg, excluding interferences caused by other materials in
      the mud.
1.6   Any modification of this method beyond those expressly permitted shall be considered as a
      major modification subject to application and approval of alternative test procedures under
      40 CFR 136.4 and 136.5.
1.7   The gas chromatography portions of this method are restricted  to use by or under the
      supervision of analysts experienced  in the use of gas chromatography and in the interpretation
      of gas chromatograms. Each laboratory that uses this method must generate acceptable results
      using the procedures described in Sections 7.1, 8.2, and 12 of  this method.

2.    SUMMARY OF METHOD

2.1   An oil sample is diluted in hexane.  An  internal standard is added and an aliquot is injected
      into a gas chromatograph (GC). The components of the oil are separated by the GC and
      detected by a flame-ionization detector (FID).
2.2   Identification of diesel oil or crude oil (qualitative analysis) is performed by comparing  ratios
      of groups of n-alkanes.
2.3   Quantitative analysis is performed by calibrating the GC/FID with hexadecane using an
      internal standard technique. The calibration factor is then used to determine the amounts of
      the groups  of n-alkanes.  A quotient of these amounts establishes that the oil is diesel or crude.
2.4   Quality  is assured through reproducible calibration and testing  of the extraction and GC
      systems.
                                                                                          57

-------
Method  1663
3.     CONTAMINATION AND INTERFERENCES

3.1    Solvents, reagents,  glassware, and other sample processing hardware may yield artifacts and/or
       elevated baselines causing misinterpretation of chromatograms.
       3.1.1   All materials used in the analysis shall be demonstrated to be free from interferences
               by running method blanks initially  and with a sample batch (samples started through
               the extraction process at the same time, to a maximum of ten).  Specific selection of
               reagents and purification of solvents by distillation in all-glass systems may be
               required.
       3.1.2   Glassware  and, where possible, reagents are cleaned by rinsing with solvent or baking
               at 450°C for a minimum of I hour.
3.2    Interferences co-extracted from samples  may vary from  source to source, depending on the
       diversity of the site being sampled.

4.     SAFETY

4.1    The toxicity or carcinogenicity of each reagent used in this method has not been defined.
       Therefore, each chemical compound should be treated as a potential health hazard. From this
       viewpoint, exposure to these chemicals must be reduced to the lowest possible level by
       whatever means available.
4.2    The laboratory is responsible for maintaining a current awareness file of OSHA regulations
       regarding the safe handling of the chemicals specified in this  method.  A  reference file of
       material safety data sheets (MSDSs) should  also be made available to all personnel involved in
       the chemical  analysis.  Additional references to laboratory safety can be found in References I
       through 3.

5.     A PPA RA TUS A ND MA TERIA L S

       NOTE:  Brand names, suppliers, and pan numbers are for illustrative purposes only.
       No endorsement is implied.  Equivalent performance may be achieved using apparatus
       and materials other than those specified here, but demonstration of equivalent
       performance meeting the requirements of this method is  the responsibility of the
       laboratory.

5.1    Equipment for glassware cleaning.
       5.1.1   Laboratory sink with overhead fume hood.
       5.1.2   Kiln: Capable of reaching 450°C within 2 hours and holding  450°C within ± 10°C,
               with temperature controller and safety switch (Cress  Manufacturing Co., Santa Fe
               Springs, CA, B31H or X31TS,  or equivalent).
5.2    Equipment for sample preparation.
       5.2.1   Laboratory fume hood.
       5.2.2   Analytical  balance:  Capable of weighing O.I mg.
       5.2.3   Glassware.
58

-------
                                                                                   Method 1663
               5.2.3.1   Disposable pipettes:  Pasteur, 150 mm long by 5 mm i.d. (Fisher Scientific
                         13-678-6A, or equivalent).
               5.2.3.2   Glass pipettes:  0.1-, 1.0-, and 10-mL, accurate to 1% or better.
               5.2.3.3   Volumetric flasks: Glass, 10- and 100-mL.
               5.2.3.4   Sample vials:  Amber glass, 1- to 3-mL with PTFE-lined screw- or crimp-
                         cap, to fit GC autosampler.
5.3    Gas Chromatograph (GC): Analytical system with split injection, capillary column,
       temperature program with initial and final isothermal holds, and all required  accessories,
       including syringes, analytical columns, gases, detector, and recorder.  The analytical system
       shall meet the performance specifications in Section 12.
       5.3.1   Column:  30 m long (+5 m) by 0.25 mm i.d. (±0.02 mm), 99% methyl, 1% vinyl,
               1.0 pm film thickness, bonded-phase fused-silica capillary (Supelco  SPB-1, or
               equivalent).
       5.3.2   Detector:  Flame ionization.  Capable of detecting 10 ng of hexadecane.
5.4    GC data system:  Shall collect and record GC data, store GC runs in magnetic memory or on
       magnetic disk or tape, process GC data, compute peak areas, store calibration data including
       retention times and the response factor, identify GC peaks through retention  times, and
       compute concentrations.
       5.4.1   Data acquisition:  GC data shall be collected continuously throughout the analysis and
               stored on a magnetic storage device.
       5.4.2   Response factor: The data system shall be used to record and maintain the response
               factor  (Section 7). Computations of relative standard deviation (coefficient of vari-
               ation;  CV) are used for testing calibration linearity.  Statistics on initial (Section 8.2)
               and ongoing (Section 12.5) performance shall be computed and maintained.
       5.4.3   Data processing: The data system shall search, locate, identify, and quantify the
               compounds of interest in each GC analysis.  Software routines shall  be employed to
               compute and record retention times and peak areas.  Displays of chromatograms and
               library comparisons are required to verify results.

6.     REAGENTS

6.1    Hexane:  ACS  grade or equivalent.
6.2    Standards: Purchased as solutions or mixtures with certification as to their purity,
       concentration, and authenticity, or prepared from materials of known purity  and composition.
       If compound purity is 96% or greater, the weight may be used without correction to compute
       the concentration of the standard.  If diesel oil in drilling mud is to be tested, the diesel oil
       standard used in this method should be from the diesel oil added to the mud  on the drilling rig
       from which the mud sample is taken.   If this oil is not available, No. 2 diesel oil from  a local
       source may be  substituted.
       6.2.1   Stock solutions:  Prepare in hexane for injection into the GC.  Observe the safety
               precautions in  Section 4.

-------
Method  1663
               6.2.1.1   Diesel oil (62.5 mg/mL):  If QC extracts from Method 1662 are to be
                         tested, use the oil that was spiked to produce these extracts. Weigh 6.25 g
                         of diesel oil to three significant figures in a 100-mL ground-glass stoppered
                         volumetric flask and fill to the mark with hexane.  After the oil is
                         completely dissolved, transfer the solution to a  150-mL bottle with
                         PTFE-lined cap.
               6.2.1.2   Normal hydrocarbons—decane (C,2), hexadecane (C16), and tetracosane
                         (C24):  Dissolve an appropriate amount of reference material in a suitable
                         solvent.  For example, weigh 10.0 mg of decane in a 10-mL volumetric
                         flask and fill to the mark with hexane. After the decane is completely
                         dissolved, transfer the solution to a 15-mL vial  with PTFE-lined cap.
               6.2.1.3   Internal standard:  Dissolve  1.0 g of 1,3,5-trichlorobenzene (TCB, Kodak
                         No.  1801  or equivalent) in 100  mL hexane.  After the TCB is completely
                         dissolved, transfer the solution to a 150-mL  bottle with PTFE-lined cap.
                         Label with the concentration and date. Mark the level of the meniscus on
                         the bottle to detect solvent loss.
               6.2.1.4   Stock solutions should be  checked  for signs of degradation prior to the
                         preparation of calibration  or performance test standards.
       6.2.2   Normal hydrocarbon calibration standards (CAL): Dilute and mix the stock solutions
               (Section 6.2.1.2) in hexane to produce the  calibration standards shown in Table 1.
               The three solutions permit the response of hexadecane to be measured  as a function of
               concentration, and decane and tetracosane permit the retention-time window of diesel
               oil to be defined.  The medium-level  solution  is used for calibration verification.
       6.2.3   Precision and recovery standard: Dilute the stock solution of diesel oil
               (Section 6.2.1.1) to produce a concentration of 1.25 mg/mL in hexane. This standard
               is used for initial precision and recovery (IPR, Section 8.2) and ongoing precision and
               recovery (OPR, Section 12.5).
       6.2.4   Addition of internal standard:   Using a  micropipette or microsyringe, transfer 100 /iL
               of each standard solution (Section 6.2.2 or 6.2.3) to a GC  injection vial. Add  100 pL
               of the TCB internal standard (Section 6.2.1.3) to each vial  and mix thoroughly.
               Calibration and precision and  recovery  standards are made fresh daily to avoid solvent
               loss by evaporation.
       6.2.5   Stability of standards.
               6.2.5.1   When not being used, standards are stored in the dark at -20  to  -10°C in
                         screw-capped vials with PTFE-lined lids.  A mark  is placed on the vial at
                         the level of the solution so that  solvent loss by evaporation can be
                         detected.  The vial is brought to room temperature prior to use.  Any
                         precipitate is redissolved and solvent is added if solvent loss has occurred.
               6.2.5.2   Standard solutions used  for quantitative purposes (Sections 6.2.1 through
                         6.2.3)  shall  be analyzed within 48  hours of preparation and on a monthly
                         basis thereafter for signs of degradation.  Standards will remain acceptable
                         if the peak area remains within  + 15% of the area obtained in the initial
                         analysis of the standard.

60

-------
                                                                                    Method 1663
7.     CALIBRATION

7.1    Establish gas chromatographic operating conditions given in Table 2.  Verify that the GC
       meets the minimum level in Table 4.  The gas chromatograph is calibrated using the internal
_ standard technique. _
       NOTE: Because each GC is slightly different, it may be necessary to adjust the
       operating conditions (carrier gas flow rate and column temperature and temperature
       program) slightly until the retention times in Table 3 are met.

7.2    Internal standard calibration procedure:  1,3,5-Trichlorobenzene (TCB) has been shown to be
       free of interferences from the diesel and crude oils tested in the development of this method.
       Check for acceptability by injecting 0.5 /xL of the internal standard solution (Section 6.2.1.3)
       into the GC\FID.  If a major peak other than the TCB peak appears in the chromatogram,
       interference with the peaks  used for determination of diesel/crude oil  may occur.  In this case,
       the analyst must choose  an  alternative internal standard that is free from interferences.
       7.2.1   Inject 1  /zL of each calibration standard containing the internal standard (Table 1 and
               Section 6.2.2) into the GC\FID.   The TCB will elute approx 8.5 minutes after
               injection.  For the GC\FID used in the development of this method, the TCB internal
               standard peak was  30 to 50% of full scale at an attenuator setting of 8 x 10"11 amp.
       7.2.2  Response  factor of hexadecane (C16).
               7.2.2.1    Tabulate the peak areas against concentration for  the TCB and C16 peaks.
                          Calculate response factors (RF) at each concentration  for C16 using the
                          following equation:


                                           Equation 1

                                          RF   (
                    where:
                     As = Area of the peak to be measured
                    Ch = Concentration of the internal standard, in pglkg
                    A^ = Area of the internal standard peak
                     C. = Concentration of the peak to be measured, in
               7.2.2.2   Calculate the average, standard deviation, and relative standard deviation
                          (RSD) of the response factors.  If the RF is constant (< 15%  RSD) over
                          the calibration range, linearity through the  origin can be assumed and
                          system performance is acceptable; if not, the system must be  recalibrated.
               7.2.2.3   The average response factor is verified on each working 8-hour shift by
                          measurement of the medium-level calibration standard (Section 12.4).
       7.2.3   Single-point calibration for  diesel oil:  Inject the precision and recovery standard
               (Section 6.2.3) to which the internal standard has been added (Section 6.2.4) to
               produce a single calibration point for diesel oil.
                                                                                              61

-------
Method 1663
               7.2.3.1   Integrate the area of the C16 peak.
               7.2.3.2   Determine the response factor for diesel oil using Equation 1.

8.     QUALITY ASSURANCE/QUALITY CONTROL

8.1    Each laboratory that uses this method is required to operate a formal quality assurance
       program (Reference 4).  The minimum requirements of this program consist of an initial
       demonstration of laboratory capability, an ongoing analysis of standards and blanks as a test of
       continued performance, analyses of spiked samples to assess accuracy,  and analysis of
       duplicates to assess precision. Laboratory performance is compared  to established
       performance criteria to determine if the results of analyses meet the performance
       characteristics of the method. If the determination of diesel/crude oil is to be made on extracts
       from Method 1662,  the quality control samples for initial precision and recovery (IPR),  spiked
       samples, duplicates, and ongoing precision and recovery (OPR) from Method 1662 shall be
       substituted for those in the QC tests below, and the specifications in Table 4 of this method for
       extracts from Method  1662 shall be met.
       8.1.1   The analyst shall make an initial demonstration of the ability to generate acceptable
               accuracy and  precision with this method.  This ability is established as described in
               Section 8.2.
       8.1.2   The analyst is permitted to modify this method to improve separations or lower the
               costs of measurements,  provided  all performance requirements are met.  Each time a
               modification is made to the method, the analyst is required to achieve the minimum
               level (Section 7.1) and to repeat the procedure in Section 8.2 to demonstrate method
               performance.
       8.1.3   Analyses of spiked samples are required to demonstrate method accuracy when
               extracts from Method 1662 are analyzed.  The procedure and QC criteria for spiking
               are described in Section 8.3.
       8.1.4   Analyses of duplicate samples are required to demonstrate method precision when
               extracts from Method 1662 are analyzed.  The procedure and QC criteria for
               duplicates are described in Section 8.4.
       8.1.5   Analyses of blanks are required to demonstrate freedom from contamination.  The
               procedures  and criteria for analysis of a blank are described in  Section 8.5.
       8.1.6   The laboratory shall, on an ongoing basis, demonstrate through calibration verification
               and the analysis of the precision and recovery standard (Section 6.2.3) that the
               analysis system is in control.  These procedures  are described  in Section 12.
       8.1.7   The laboratory shall maintain records to define the quality of data that is generated.
               Development of accuracy statements is described in Sections 8.3.2 and 12.5.4.
8.2    Initial precision and accuracy:  The initial  precision and recovery test is performed using the
       precision and recovery standard.  If extracts from Method  1662 are to be analyzed, the
       extracts from the initial precision and recovery test in that method shall  be used; otherwise, the
       laboratory shall generate acceptable precision and recovery by performing the following
       operations.

-------
                                                                                    Method 1663
       8.2.1   Using diesel oil, prepare four separate aliquots of the precision and recovery standard
               (Section 6.2.3) using the procedure in Section 10. Add the internal standard to each
               aliquot (Section 6.2.4). Analyze these aliquots using the procedure in Section 11.
       8.2.2   Using results of the set of four analyses, compute the average recovery  (X)  in mg/mL
               and the standard deviation of the recovery (s) in mg/mL for each sample by the
               internal standard method (Sections 7.2 and 14.2).
       8.2.3   For each compound, compare s and X with the corresponding limits for initial
               precision and accuracy in Table 4.  If s and X meet the acceptance criteria,  system
               performance is acceptable and analysis of samples may begin.  If, however, s exceeds
               the precision limit or X falls outside the range for accuracy, system performance  is
               unacceptable. In this event, review this method, correct the problem, and repeat  the
               test.
8.3    Method accuracy:  If extracts from Method 1662 are to be analyzed, the extract  from the
       accuracy test in that method  shall be used; otherwise, an accuracy  test is unnecessary.  The
       procedure for determining method accuracy is given in Section 8.3 of Method 1662, and the
       specification for accuracy is  given in Table 4 of this method.
       8.3.1   Compare the percent recovery for diesel oil with the corresponding QC acceptance
               criteria in Table 4.  If the results of the spike fail the acceptance criteria, and the
               recovery of the QC  standard in the ongoing precision and recovery test  (Section
               12.6.3) is within the acceptance criteria in Table 4, an interference may be present.
               In this case,  the result may  not be reported for regulatory compliance purposes.  If,
               however, the results of both the spike and the ongoing precision and recovery test fail
               the acceptance criteria, the  analytical system is judged to be out of control, and the
               problem must be immediately identified and corrected and the sample batch
               reanalyzed.
       8.3.2   As part of the QA program for the laboratory,  method accuracy for samples shall be
               assessed and records shall be maintained.  After the  analysis of five spiked samples
               in which the recovery passes the test in Section 8.3.1, compute the average percent
               recovery (P) and the standard deviation of the percent recovery (sp).  Express the
               accuracy assessment as a percent recovery interval from P  2sp to P +  2sp.  For
               example, if P = 90% and sp =  10% for five analyses of diesel oil, the  accuracy
               interval is expressed as 70 to 110%.  Update the accuracy assessment on a regular
               basis (e.g., after each five to ten new accuracy measurements).
8.4    Duplicates:  If extracts from Method 1662 are to be analyzed, the  extracts from  the duplicates
       test in that method shall  be used.  The procedure for preparing duplicates is given in Section
       8.4 of Method 1662, and the specification for RPD is given in Table 4 of this method. If
       extracts from Method 1662 are not to be analyzed, duplicates  of the precision and recovery
       standard (Section 6.2.3)  are analyzed, and the specification for RPD is given for diesel oil in
       Table 4 of this method.
       8.4.1   Analyze each of the duplicates per the procedure in Section 11  and compute the
               results per Section 14.
                                                                                              63

-------
Method 1663
       8.4.2   Calculate the relative percent difference (RPD) between the two results per the
               following equation:


                                          Equation 2

                                   RPD =  '   ' " _ i_ x 100
                 vv/zere:
                 D,  = Concentration of diesel oil in the sample
                 D2  = Concentration of diesel oil in the second (duplicate) sample
       8.4.3   The relative percent difference for duplicates shall meet the acceptance criteria in
               Table 4.  If the criteria are not met, the analytical system shall be judged to be out of
               control, and the problem must be immediately identified and corrected and the sample
               batch  reanalyzed.
8.5    Blanks:  If extracts from Method 1662 are to be analyzed, the extracts from blanks in that
       method shall be analyzed in addition to the blanks in this method.
       8.5.1   Rinse the glassware used in preparation of the extracts  in this method with hexane and
               analyze a 1-^L aliquot of the rinsate using the procedure in Section 11.  Compute the
               results per Section  14.
       8.5.2   If any peak is detected in a blank at greater than the minimum level in Table 1,
               analysis of samples is halted until the  source of contamination is eliminated and a
               blank  shows no evidence of contamination.
8.6    The specifications contained in this method can be met if the apparatus used is  calibrated
       properly, then  maintained in a calibrated state.  The standards used for initial precision and
       recovery (IPR, Section 8.2) and ongoing precision and recovery (OPR, Section  12.5) precision
       and recovery should be identical, so that the most precise results will be obtained.  The GC
       instrument will provide the most reproducible results if dedicated to the settings and conditions
       required for the analyses given in this method.
8.7    Depending on specific program requirements, field replicates  and field spikes of diesel oil into
       samples  may be required when Method  1662 and this  method are used to assess the precision
       and accuracy of the  sampling and sample transporting techniques.

9.     SAMPLE  COLLECTION, PRESERVATION, AND HANDLING

9.1    Oil samples are collected in 20- to 40-mL vials with PTFE- or aluminum-foil-lined caps and
       stored in the dark at -20 to -10°C.
9.2    If extracts  from Method 1662 are to be analyzed, the laboratory should be aware that sample
       and extract holding times for this method have not yet been established.  However, based on
       tests of wastewater for the analytes determined  in this method, samples shall be extracted
       within seven days of collection and  extracts shall be analyzed within 40 days of extraction.
64

-------
                                                                                 Method 1663
9.3   As a precaution against analyte and solvent loss or degradation, sample extracts are stored in
      glass bottles with PTFE-lined caps, in the dark, at -20 to -10°C.

10.  DILUTION OF OIL AND EXTRACTS

10.1  Neat oil samples:  If oil is received in neat form,  it should be diluted to bring the
      concentration within the range of the instrument.  If the oil is No. 2 diesel oil, the appropriate
      concentration will be approximately  1.25 mg/mL.
10.2 Extracts from Method 1662:  If extracts of samples from Method 1662 are to be analyzed,
      these extracts (from Section 10.4.2 of that method) are analyzed undiluted unless diesel oil  is
      known or suspected to be present.
10.3 Neat oil expected to be diesel oil.
      10.3.1 Weigh 100 mg into a 10-mL volumetric flask and dilute to the mark with hexane to
              produce  a concentration of 10 mg/mL. Stopper and mix thoroughly.
      10.3.2 Using a calibrated  1.0-mL volumetric pipette, withdraw 1.0-mL of the solution
              created in Section  10.3.1.1 and place in a 10-mL volumetric flask.  Then withdraw an
              additional 0.25 mL of the solution and add it to the  10-mL volumetric flask (for a
              total of 1.25 mL).   Fill to the mark with hexane to  produce a concentration of 1.25
              mg/mL (1250 /ig/mL). This solution will be near, but not above, the limit of the
              calibration range and will match the concentration of the QC samples from
              Method  1662 (assuming 100% recovery).

7 7.  GAS CHROMA  TOGRAPHY

11.1  Table 2 summarizes the recommended operating conditions for the GC.  Retention times for
      the n-alkanes obtained under these conditions are  given in Table 3.  An example of the
      separation achieved for diesel oil is shown in Figure  1.  Other columns, chromatographic
      conditions, or detectors may be used if the minimum level (Section 7.1) and the initial
      precision  and accuracy  requirements (Section 8.2) are met.
11.2 Using a micropipette or microsyringe, transfer equal  100-/iL volumes of the sample, sample
      extract, or QC standard extract (Section 10.2) and the TCB internal standard solution
      (Section 6.2.1.3)  into a GC injection vial.  Cap tightly and mix thoroughly.
11.3 Inject 1 tiL of the sample extract or standard into  the GC,  using the conditions in Table 2.
11.4 Begin data collection and the temperature  program at the time of injection.
11.5 If the area of any peak exceeds the calibration range of the system, dilute a fresh aliquot of the
      extract by a factor of 10, mix 100 pL of internal  standard with a  100-juL aliquot of the extract,
      and reanalyze.
11.6 Compute the concentrations of the individual n-alkane peaks  using the response factor for
      hexadecane from  the calibration data (Section 7.2.2.2).

12.  SYSTEM AND LABORA TORY PERFORMANCE

12.1  At the beginning  of each 8-hour shift during which analyses  are performed,  GC calibration and
      system performance are verified. For these tests, analysis of the medium-level calibration

                                                                                           65

-------
Method 1663
       standard (Table 1) and of the precision and recovery standard (Section 6.2.3) shall be used to
       verify all performance criteria.  Adjustment and/or re-calibration (per Section 7) shall be
       performed until all performance criteria are met.  Only after all performance criteria are met
       may samples and blanks be analyzed.
12.2  Inject 1 /*L of the medium-level calibration standard (Table 1) into the GC  instrument
       according to the procedure in Section 11.
12.3  Retention times.
       12.3.1  Retention time of the internal standard:  The absolute retention  time of the TCB
               internal standard shall be within the range of 7.96 to 8.08 minutes.
       12.3.2  Relative retention times of the n-alkanes:  The retention times of the n-alkanes relative
               to the TCB internal standard shall be within the limits given in  Table 3.
12.4  Calibration verification:  Compute the concentration of hexadecane based on the average
       calibration factor (Section 7.2.2.2).  The concentration shall be within the limits in Table 4.  If
       calibration is verified, system performance is acceptable and analysis of blanks and QC
       samples  may begin.  If, however, the concentration falls outside of the calibration verification
       range, system performance  is unacceptable.   In this case, correct the problem and repeat the
       test, or recalibrate (Section 7).
12.5  Ongoing precision and recovery (OPR):  If the extract is from Method  1662, the OPR standard
       from that method shall be used and the specification for the OPR from Method  1662 in Table
       4 shall be met;  if not, a sample of diesel oil shall be diluted per the procedure in Section 10
       and shall be used for the OPR test.
       12.5.1  Analyze the appropriate OPR standard.
       12.5.2  Compute the concentration of diesel oil  in this standard per Section 14.2.
       12.5.3  Compare the concentration with the limits for ongoing precision and recovery in
               Table 4.  If the concentration is in the range specified, the analytical processes are in
               control and analysis of blanks and samples may proceed.  If, however, the
               concentration is not in the specified range, these processes are not in control.  In this
               event, correct the problem,  re-extract the sample  batch if the OPR is from
               Method 1662, or redilute the oil sample (per Section 10.3) and  repeat the ongoing
               precision and recovery test.
       12.5.4  Add results that pass the specifications in Section 12.5.3 to initial and previous
               ongoing data.  Update QC charts to form a graphic representation  of continued
               laboratory performance.   Develop statements of laboratory data quality for diesel oil
               by calculating the average percent recovery (R) and the standard deviation of percent
               recovery (sr). Express the accuracy statement as  a recovery interval from R   2sr to
               R +  2sr.  For example, if R = 95% and sr = 5%, the accuracy is 85 to 105%.

13.   QUALITATIVE IDENTIFICATION

13.1   Qualitative identification is accomplished by comparison of data from analysis of a sample or
       blank with data from analysis of the calibration verification standard (Section 12.4).  Diesel
       and crude oil are differentiated by the presence and concentration of the Cg-Cjo n-alkane peaks
       in the chromatogram of extracts of the sample.
66

-------
                                                                                  Method 1663
13.1  Using the calibration data, establish the identity of the C9-Cy, n-alkane peaks in the
      chromatogram of the sample.
13.2  Diesel oil is not present in the sample if there are less than 10 n-alkane peaks present in the
      C9-C24 range at  a signal-to-noise ratio equal to or greater than 3 for each peak, and if the QC
      tests (Sections 8 and 12) for the sample set are acceptable.  The experience of the analyst shall
      weigh heavily in the determination of the presence of peaks at a signal-to-noise ratio of 3  or
      greater.
13.3  If ten or more n-alkane peaks are present in the analysis of the sample, diesel oil, mineral oil,
      or crude oil may be present.  Mineral oil can be distinguished by its lower polynuclear
      aromatic hydrocarbon content using Method  1654A.  Some crude oils may be distinguished by
      the presence and concentration of n-alkanes in the C^-C^ range. If peaks are present in the
      Cjs-C^ range, the quantitative measurements  in Section 14 are used as a final determination
      that the oil is crude oil.

14.  QUANTITATIVE DETERMINATION

14.1  Differentiation between diesel and crude oil.
      14.1.1   Using the concentrations of the individual n-alkane peaks determined in
               Section 11.6, sum the concentrations of the n-alkanes from C9 to C^, inclusive.
               Similarly, sum the concentrations of the n-alkanes from C& to C^, inclusive.
      14.1.2  Calculate the percentage  of C^-C^ n-alkanes as follows:


                                          Equation  3
                                          Sum of €.,-€„. n-alkanes
                      Percent^ - CJ =	J*   *	x 100
                                          Sum of Cg-Cy, n-alkanes

      14.1.3  If the percent of C^-C^  n-alkanes is greater than  1.2, the oil is  crude oil.
14.2  Determination of diesel oil:  Compute the concentration of diesel oil in the standard or QC
      extract using the hexadecane peak only,  and the response factor given in  Section 7.2.2.2, using
      the following equation:
                                                                                            67

-------
Method 1663
                                        Equation 4

                                   Crx (mg/mL) = _£
                    where:
                    Ca = Concentration of oil in the sample
                    C.u = Concentration of the internal standard, in mg/mL
                     As = Area of the peak to be measured
                    A^ = Area of the internal standard peak

                    (For RF, see Equation 1)
15.   COMPLEX SAMPLES

15.1   The most common interferences in the determination of diesel oil are from mineral oil and
       proprietary additives in drilling fluids, and from naturally occurring hydrocarbons from crude-
       oil-bearing formations.
15.2   Mineral oil can be identified by its lower polynuclear aromatic hydrocarbon content using
       Method 1654 A.
15.3   Crude oils can usually be distinguished by the percentage of C^-CM n-alkanes per Section 14
       of this method. However, some crude oils may not produce peaks  in the C^-C^ range.
       15.3.1 Oil condensates from gas wells  are low in molecular weight and will normally
              produce chromatographic peaks in the C8-C16 range.  If a sample of the gas
              condensate crude oil from the formation is available, the oil can be distinguished from
              diesel oil using the extract from this method and the n-alkane ratio determinations in
              the section on qualitative determination in Method  1651.
       15.3.2 Asphaltene crude oils with API gravities <20 may not produce chromatographic
              peaks in the C^-C^ range.  In  this instance, the lack of peaks  in the C^-Cy, range
              cannot be used to prove that the oil is crude oil and not diesel oil.  However, the
              absence of ten peaks in the C9-C24 range can be used to demonstrate that diesel oil is
              not present, per Section 13 of this method.

16.   METHOD  PERFORMANCE
Specifications in this method are adopted from EPA Method  1651 (Reference 5).  Example
chromatograms of diesel  oil and crude oil are shown  in Figure 1.
68

-------
                                                                              Method 1663
                                    References
1.     "Carcinogens—Working With Carcinogens."  Department of Health, Education, and Welfare,
      Public Health Service, Centers for Disease Control [available through National Technical
      Information System, 5285 Port Royal Road, Springfield, VA 22161, document no. PB-
      277256]: August 1977.

2.     "OSHA Safety and Health Standards, General Industry [29 CFR 1910], Revised."
      Occupational Safety and Health Administration, OSHA 2206. Washington, DC: January 1976.

3.     "Safety in Academic Chemistry Laboratories (3rd Edition)."  American Chemical Society
      Publication, Committee on Chemical Safety.  Washington, DC: 1979.

4.     "Handbook of Analytical Quality Control in Water and Wastewater Laboratories."  USEPA,
      EMSL-Ci, EPA-600/4-79-019. Cincinnati, OH: March 1979.

5.     "Method 1651, Total Oil and Diesel Oil in Drilling Muds and Drill Cuttings by Retort,
      Gravimetry, and GC/FID."   Available from the EPA Sample Control Center, 300 N. Lee St.,
      Alexandria, VA  22314.
                                                                                       69

-------
Method 1663
Table 1.  Concentration of Calibration Standards

                                  Calibration Solution Concentration fug/mLJ

            Ana/yte

            n-decane
            n-hexadecane
            n-tetracosane
            Diesel Oil
Low
--
10
—
—
Medium
40
40
100
1250
High
—
200
—
--
Table 2.  Gas Chromatographic Operating Conditions

Injection port, transfer fine, and detector temperatures: 275°C
Column temperature program:
      Initial temperature: 90°C
      Initial time: 0 minutes
      Ramp: 90°C-250°C @ 5°C per minute
      Final temperature:  250°C
      Final hold: 10 minutes or until all peaks have eluted
Carrier gas and flow rates:
      Carrier: Nitrogen or helium
      Velocity: 20-40 cm/sec @ 90°C
      Split ratio:  0-120:1**
      Makeup gas: As required by manufacturer
Hydrogen and air flow rates: As specified by manufacturer
Detector amplifier settings: 10"n amp full scale. Attenuation is adjusted so that the highest
      peaks are on scale in the most concentrated standard.
Recorder:  Chart speed of 1 -2 cm/min (fixed)

* Conditions are approximate and can be adjusted to meet the performance criteria in Section 12
  (see the note in Section 7.1).
**Lower split ratios may give more reproducible results.
70

-------
                                                                            Method 1663
Table 3.  Retention Times and Relative Retention Time Limits for n-Alkane Peaks
                                           Retention Time
                     Compound

                     TCB
                     n-C12
                     n-C13
                     n-Cu
                     n-C16
                     n-C16
                     n-C17
                     n-C18
                     n-C19
                     n-C20
                     n-C21
                     n-C22
                     n-C23
                     n-C24
Mean
(minutes)
8.0
9.9
12.6
15.3
17.9
20.4
22.9
25.2
27.3
29.4
31.5
33.4
35.3
37.1

Relative
1.00-1.00
1.22-1.24
1.55-1.57
1.89-1.92
2.21-2.25
2.52-2.56
2.82-2.88
3.12-3.15
3.39-3.43
3.66-3.71
3.90-3.97
4.14-4.21
4.37-4.45
4.58-4.69
Table  4.  QC Acceptance Criteria
Criterion                       Units

Minimum Level2                fjg/mL
Method Detection Limit3         mg/kg
Initial Precision and Recovery
   Precision (RSD)4              mg/mL
   Recovery                    mg/mL
Calibration Verification5          pg/mL
Ongoing Precision and Recovery4 mg/mL
Matrix Spike Recovery4            pet
Duplicates                      RPD
Diesel
 Oil1

 100
 100
Oil from Method
     1662
                  0.23
               1.00-1.35

               0.98-1.37
               0.73-1.14
                   16
n-hexadecane
      10
                       34-46
1  CAS Registry number 68534-30-5; #2 diesel oil used for these tests.
2  This is a minimum level at which the analytical system  shall give recognizable signals and
   acceptable calibration points.
3  Estimated; 40 CFR Part 136, Appendix B; MDL is diesel oil in mud.
4  Test concentration in diluted extract = 1.25 mg/mL
5  Test concentration = 40 //g/mL
                                                                                     71

-------
Method 1663
Crude Oil
                O
 in
 i—

O
                                          l/sl
               O
               CM

              O
Diesel Oil
                                                                                 A 52-010-003
Figure 1.  Chromatograms of Crude Oil and  Diesel Oil, Showing Differences in the C25-C30

          Range
72

-------