EPA-AA-IMS/80-5-C
      TECHNICAL REPORT
          MARCH 1981
 RECOMMENDED SPECIFICATIONS
             FOR
EMISSION INSPECTION ANALYZERS:
   CHANGE NOTICE NUMBER 1

-------
                                                             EPA-AA-IMS/80-5-C
                                Technical Report
                                   March 1981
                           Recommended Specifications
                                      For
                         Emission Inspection Analyzers:
                             Change Notice  Number 1
                                       By

                              William B. Clemmens
                                     Notice
This Report  does  not necessarily represent  final  EPA decisions  or  positions.
It is intended to present technical analysis of  the issue using  data  which  are
currently available.  The purpose  in the release of  such reports is  to facil-
itate  the  exchange  of  technical  information  and  to inform   the  public   of
technical developments  which  may  form  the basis  for a  final   EPA  decision,
position or regulatory action.


NOTE:  This  report  is  the  third in a  series  of reports.   Report  EPA-AA-IMS/
80-5-A contains  background,  technical discussions,  and  policy  information  on
Inspection  Analyzers.   Report  EPA-AA-IMS/80-5-B   provides  the  Recommended
Technical Specifications  for inspection  analyzers.   This report  (EPA-AA-IMS/
80-5-C)  provides  modifications  to   those   specifications.   The  reports  are
available separately.


                        Inspection and Maintenance Staff
                      Emission Control Technology Division
                  Office of Mobile Source Air Pollution Control
                      Office of Air, Noise,  and  Radiation
                      U.S. Environmental Protection Agency

-------
                               TABLE OF CONTENTS
I.    Forward                                                            3


II.   Modified Sections in Chapter VI

     A.   Section D:   Definitions and Abbreviations                      5


III.  Modified Sections in Chapter VII

     A.   Recommended Qualification Program                              6
     B.   Section A.  Gases                                               8
     C.   Section B.  Gas Cylinders                                       9
     D.   Section D.  Design Requirements                                10
     E.   Section E.  Analyzer Performance Specifications                 13
     F.   Section F.  Sample System Performance Specifications            16
     G.   Section G.  Operating Environment                              18
     H.   Section H.  Fail-Safe Features                                 19


IV.   Modified Sections in Chapter VIII

     A.   Introduction to Chapter VIII                                  20
     B.   Section A.  Automatic Zero/Span Check                          21
     C.   Section B.  Automatic Leak Check                               22
     D.   Section E.  Dual Tailpipes                                     23
     E.   Section H.  Vehicle Diagnosis                                  24
     F.   Section I.  Anti-Tampering                                     25


V.    Modified Sections in Chapter IX

     A.   Section B.                                                     26
     B.   Section C.  Anti-Dilution                                      27
     C.   Section D.  Loaded Mode Kit                                    28


VI.   Modified Sections in Chapter XI

     A.   Introduction to Chapter XI                                    29
     B.   Specific Test Procedure Changes                               31

-------
I.  FORWARD

    In September of 1980, EPA published  two  technical  reports dealing with I/M
inspection analyzers.  The first report,  "Analysis  of  the Inspection Analyzer"
(EPA-AA-IMS/80-5-A),  discussed  background information  on the  analyzers  them-
selves, provided a brief comparison of  analyzers  specifications,  listed  some
sample cost calculations, and indicated  a few policy implications.   The second
report,   "Recommended   Specifications   for   Emission   Inspection   Analyzers"
(EPA-AA-IMS/80-5-B) detailed  the  I/M Staff's  findings on  analyzer specifica-
tions for both a manually operated  inspection analyzer and  a computer operated
inspection analyzer.

    Once EPA's  analyzer  recommendations  were  formalized,  analyzer  manufactur-
ers could  then  justify  the time and  expense of  initiating  engineering design
studies targeted  towards building  an  EPA analyzer.   From these studies  came
several  requests  for   better  interpretation   or   clarification-  of  certain
portions  of  the   EPA  specifications.    In   other  cases,  the  specifications
stimulated  the  manufacturers  to come  up with  "better mouse  trap" approaches
that  met   the  intent  of  the  specifications,  but  might  not have  passed  a
bureaucratic interpretation of  the  specifications.   In still  other areas,  the
manufacturers  offered  new  evidence  or  more  convincing  arguments  that  a
relaxation  of  the specifications  in a  few  specific  instances was  necessary
from a technical perspective in relation to the targeted market and cost.

    As  discussed  in  the  "Recommended  Specifications   ..."  (80-5-B),   the
technical  report  format  used  by the  Office  of Mobile  Source Air  Pollution
Control (OMSAPC) has no  formal  mechanism for instituting a  change  to previous
reports other than publishing  a new report.   This  report (80-5-C)  constitutes
such  a change-notice  to  EPA-AA-IMS/80-5-B,  and  when   published,  officially
superceeds those portions of 80-5-B.

    This   report   does   not  recompile   the  original   specification  report
(80-5-B).  With only a  few exceptions,   the  format  of this  report  (80-5-C) is
laid out  in two sequences which are used interchangeably.    In all cases,  the
chapter and  section  of  80-5-B are  identified.   As  appropriate, the subsection
and item are identified  followed by a descriptive sequence of either:
                      •

    a)   Previous Concept:
    b)   New Concept:
    c)   Reason for Change:
    d)   New Wording:

or

    a)   Previous Wording:
    b)   New Wording:
    c)   Reason for Change:

In most cases,  the format keys  around  the word  "item", and  the new wording is
a direct  replacement for the  previous wording under  that "item"  (e.g. Section
D, Subsection 3.d),  Item i)2);  or  Section I,  Item, Subsection 2.).  Following
the  "item"  heading  is  the  page number  location  (in  parenthesis)  of  that

-------
heading in report 80-5-B.  The few  cases  that  are not in  this  specific format
are reasonably self-evident.

    The changes  discussed in  this  report will  not  adversely  affect  analyzer
lead-time,  the  number  of  manufacturers  interested  in  developing  an  EPA
analyzer,   or  state  implementation  schedules.   In   fact   if  anything,  these
changes (relative to 80-5-B  report) should decrease  the  lead-time  and increase
the  number  of  manufacturers  interested   in  producing   an   EPA  inspection
analyzer.   Therefore,  the publication  of these  changes  should not  cause  any
disruption in current state implementation schedules  or plans.

    EPA wishes  to  acknowledge the  valuable  comments and  considerable  efforts
put forth by the Equipment and Tool Institute  in supporting the development of
this report.

    As  always,  the  I/M staff  at  EPA's  Ann  Arbor  Facility  is  available  to
provide additional  assistance  and information as necessary.   You may  contact
Tom Cackette, Donald White, or Bill Clemmens at (313) 668-4367.

-------
II.  Modified Sections  in Chapter VI

    A.  Section D.  Defintions  and Abbreviations  (p.  13)

    1.  New Definitions

         a) Indicator,  Lights:  Indicator lights  as used  in  his document mean
         any means used to effectively  communicate  with  the  operator  (e.g.,
         lights,  computer prompts,  etc.).

         b) Output pevicg:  For  the  purpose of  this  document,  output  device
         means any  device that displays  or prints  emission values.   If  an
         output  device  displays information other  than the  emission  values
         (e.g.,  CRT  diagnostic display),  only  the  emission  value portion  of
         the multiple  display is considered the output  device.

         c) Switches;  Switches as used in  this document are  a generic term
         meaning   any  process  used  to  control  analyzer  functions   (e.g.,
         switches, buttons, alpha-numeric keyboard, etc.).

-------
                                        6

III. Modified Section in Chapter VII

    A.  Recommended Qualification Program (p. 17)

    1.  Item III. Subsequent Production QA/QC (p. 17)

         a)  Previous  Concept:   Required  full  accredidation  testing  on  all
         aspects every three years after initial certification.

         b) New  Concept:  Requires  re-certification  testing  only -on  specified
         key parameters every three years after initial certification.

         c) Reason  for Change:  The change to  recertifying  key parameters  is
         implemented to reduce the re-certification testing costs.

         d) New Wording;   The  accreditation may  be  re-certified  for a  three
         year period at any time by  passing  selected evaluation tests on  two
         of three units  selected randomly  from a production  run  of  20.   The
         selected tests will include the following:

              1.  Testing  of  all  specifications  in  Section E  of  Chapter  VII
              (Analyzer Performance Specifications).

              2.  Testing  of  all  specifications  in  Section F  of  Chapter  VII
              (Sample Systems Performance Specifications).

              3. Testing the  following  at the  high  temperature  specifications
              of  test  procedure  G.I.  in Chapter  VI  (Operating Environmental
              Test Procedures).

                   a. Analyzer Calibration Curve
                   b. Pressure and Temperature Compensation (if used)
                   c. Analyzer zero and span drift
                   d. Analyzer Response Time

              4. Testing the  following  at the  low temperature  specification of
              test  procedure  G.I.  in Chapter  XI (Operating Environment  Test
              Procedures)

                   d. System Warm-up

    2.  Item IV. QA/QC Testing Criteria (p. 18)

         a) Previous Concept: Repairs due to  random failures during  evaluation
         testing  were  allowed  on  pre-production units  only.   Justification
         required.

         b) New Concept;  Allow repair  of random  failure  during  accredidation
         and  re-certification  testing  as  long  as   justified  by  engineering
         report.

-------
c)  Reason  for  Change;   To   reduce   the  cost  of  accredidation  and
re-certification testing.

d)  New Wording:  Random  failures  must  have  sufficient  documentation
(i.e.,  published  report  available to  regulatory bodies)  to justify
why the  failure  can be attributed as a random failure  and not minor
design  failure.   Random  failures  may  be  repaired on all  units.   A
condition to allow  the repair  of production analyzers is the develop-
ment  of a  plan (where  necessary) to  prevent  the  specific  type  of
failure  in  future  production units.   After repairs,  those  tests that
might be affected by the repairs should be rerun.

-------
                                                           Spec: A.2.-3.


B. Section A. Gases (p. 19)

1. Subsection 2. Item d) zero gas

     a) Previous Wording: Zero gas may be

          i) bottled air,

          ii)  chemically  purified   room  air  such  as  with  an  activated
          charcoal trap on the anlysis system, or

          iii) catalytically purified room air.

     b)  New  Wording:  Zero  gas  may be  air  or  N£.   The   impurities   in
     bottled  zero gas  must  be analyzed at  less  than the maximum allowable
     level.   Ambient  room  air  may  be  used  for  routine   spanning   of
     inspection equipment.

     c) Reason  for  Change:  To clarify original  intent  between in-use zero
     gas and  audit gases (see sub-section 3).

2. Subsection 3. Recommended  number  of  gases:   b)  for in-use  systems. Item
ii) (p. 20).

     a) Previous Wording: purified room air  zero gas

     b)  New  Wording;   zer.o  gas may   be  drawn  from  the   ambient   air
     (room-air).  It  is  preferred that  the  pick-up  of  ambient zero air  is
     as high  off the  floor  as  practical.   Suggested accessory items (not
     required)  are  chemical  or  catalytic  devices  to  purify  the   ambient
     air,  or a  zero-air port  that  would  allow  use of ambient  zero  air
     drawn  from outside the test  cell or building.

     c)  Reason  for  Change:  To  clarify  original intent  to  not   require
     bottled  zero gas  on in-use analyzers.

3.  Subsection 3. Recommended  number of  gases c)  for  periodic  check  ...
Item iii)  (p. 20).

     a) Previous Wording: bottled zero gas

     b) New Wording;   bottled zero  gas  (because  the audit  check  will  use
     bottle  zero  gas  and  the  in-use system will  use ambient  room  air,  a
     background emission calculation must be used  during  the  audit check.)

     c) Reason  for  Change:  In allowing  ambient room  air to be  used  as zero
     gas for  in-use systems,  there can  be an offset  in  the gain setting of
     the   span  control  if  there is  a  significant  level  of background
     emissions.   Background emission  levels  can  vary  substantially from
     location to  location.   In  order   to   evaluate  the operation  of  the
     machine, background levels should be measured during the  audit  check.

-------
                                                           Spec:  B.I.

C. Section B. Gas Cylinders (p^ 21)

1. Subsection1. Item d)

     a) Previous Wording:  Disposable cylinders may not be used

     b) New Wording:  Disposable  cylinders  are generally not  desireable  in
     the  context of  gas   blending,  gas stability,  and gas  traceability.
     Also, the quantity of gas in  the disposables  is  generally quite  small
     (approximately  7-8 cubic  feet) compared  to  the  refillable  bottles
     (160 cubic  feet).  The  small  size  of the disposable  bottles  requires
     more  frequent   replacement   in the   field.    The  disposables   are,
     however,  convenient  to  handle  and   ship.    Therefore,  disposable
     cylinders may  be used .for  in-use  systems  (not  audit)  provided  that
     the suppliers of the  disposable bottles  include  the following  quality
     control measures.

          i) The  inside surfaces  of the cylinder  conform  with the  NBS CRM
          or SRM procedures for preparation,  cleanliness,  trace  materials,
          composition,  coatings,   etc.  for  the  gas  composition and  con-
          centrations used.

          ii)  The  cylinder   valve  shall  conform with  NBS  CRM  or  SRM
          procedures  for  preparation,  packing   materials,   cleanliness,
          composition,  etc.   for   the  gas  composition and  concentrations
          used.

          iii)  The  stability  over the   normal   usage  time  of  the  gas
          concentraton   shall  be  periodically   checked    from   random
          production lots by the NBS CRM procedure.

     c) Reason for_.Change:. To allow the  use of disposable cylinders.

-------
                                   10
                                                       Spec: D.3.d) & e)
                                                                 and
                                                                D.7.b)

D. Section D. Design Requirements

!• Subsection 3. Sample System d) Water Trap, Item i) 2) (p. 24)

     a) Previous  Concept;  Electronic  compensation  (based on  pressure  and
     temperature) of the water in the sample was allowed.

     b) New  Concept;  The  previous concept assumed that  the  sample leaving
     the  water  trap  was  saturated.   This  may  not  be the  case  for  all
     systems under  all  conditions.   Therefore,  the wording  for  Item 2. is
     deleted, and new wording is substituted.

     c) New Wording;

          Option  2;  Other  techniques  to  either  remove the  water  from  the
          sample  (e.g.  permapure  @,  etc.)  or correct for the water content
          (electronic dew  point,  pressure,  etc.) may be used  if  they  can
          be demonstrated to be equivalent to Option  1.

2. Subsection 3. Sample System e) Particulate Filter, Item vi) (p. 24)

     a) Previous  Concept:  The particulate filter before  the optical bench
     was  required to be 5 micron or less.

     b) New  Concept; The  particulate filter before  the optical  bench is
     preferred to be 5 micron or less.

     c) Reason for Change; To allow more flexibility  in system design.

     d) New  Wording; It is  preferred that  the  filter element  and filter
     system  shall  be designed to prevent  particulates  larger  in size than
     5  microns  from  entering  the  sample   cell  of  the   analyzer.   The
     location of  this  suggested optical bench  filter (5 micron)  is  up to
     the  manufacturer  and may  be on  the  pressure or  vacuum side  of  the
     system.  If  this filter  only filters  sample gases passing through the
     analyzer,  the manufacturer  may elect  to  have  an  additional filter
     prior to  the system pump(s).   The  particulate  size of this optional
     filter  is  at  the  discretion  of the manufacturer.  Verification of
     filter  particulate  size  removal  may  be  determined   by  the  filter
     manufacturer using standardized ASTM  or Filter Industry procedures.

3. Subsection 7.  Analyzer Spanning System,  Item b) (p. 26)

     a) Previous Wording:

     Recommended  Gas   Spanning  Frequency:  It  is   recommended  that  the
     analyzer be gas spanned  after:

          i) every  "power on" and warm-up  sequence, and

-------
                                   11
                                                       Spec:  D.7.e) & h)

          ii) every 4 hours of "power-on" condition when testing.

     b) New Wording;

     Recommended  Gas  Spanning  Frequency;   It  is  recommended  that  the
     analyzer be gas spanned after:

          i) every "power on" and warm-up sequence, and

          ii) every 4 hours of "power-on" condition when testing.

          iii) If this gas spanning  frequency  is used,  the analyzer is not
          required to have any temperature or barometric correction.

     c) Reason for Change; To clarify original intent.

4. Subsection 7. Analyzer Spanning System, Item e)  (p. 26)

     a) Previous  Wording;  Last sentence  in  subparagraph e)   ...  The audit
     port should  interconnect  the  span system downstream of  the  span/zero
     switching valve.

     b) New Wording; Delete the last sentence in subparagraph e).

     c) Reason  for Change;  The  sentence was  included  to  allow  for  leak
     checking the  span/zero  switching valve  during  the audit.   There  are
     other ways to check this valve more cost effectively.

5. Subsection 7. Analyzer Spanning System, Item h)  (p. 26)

     a) Previous  Concept;  Subparagraph  h)   required  operator  tracking  of
     the gain setting to identify analyzer problems or dirty sample cells.

     b) New  Concept:  Replaces  operator tracking by requiring  the analyzer
     manufacturers  to limit  the  overall gain adjustment  such  that  gain
     cannot  be  turned  up  to  the  point where  the  signal-to-noise ratio
     prevents the  analyzer from meeting the  calibration  curve  and drift
     requirements.

     c) Reason  for  Change;  To allow the  manufacturers  more  flexibility in
     the analyzer design.

     d) New  Wording;  (Delete the previous wording  for  subparagraph h) and
     substitute  the following).    The  overall  analyzer  span  gain-control
     shall be limited such  that  when compensating  for analyzer degradation
     the operator  runs  out of gain adjustment before  the  signal-to-noise
     ratio  of  the analyzer prevents  the analyzer  from  complying  with the
     calibration curve and drift  requirements in Section E.

-------
                                   12
                                                       Spec: D.7.1)

6. Subsection 7. Analyzer Spanning System, 1) Span Cylinders, Item i)
(p. 27).

     a) Previous Concept: Requires the  use  of  and supporting equipment for
     200 cubic foot gas cylinders.

     b)  New  Concept:  Make  specifications  consistent  with  changes  in
     Section B.  (Gas  Cylinders)  that allow  the possible  use of disposable
     cylinders.

     c) Reason for Change: Consistency with other changes.

     d) New  Wording;  For stand  alone  centralized inspection  analyzers  (a
     condition determined  by  the State  or  Contractor) and  all decentral-
     ized  analyzers,  the  analysis  system  shall  include  a  structure  for
     safely  securing  two refillable  span cylinders,  or  shall  include  an
     easily  utilized  source  for procurement  of  such  a  structure  by  the
     owner as  well as  safety and operating instructions  for use  of  the
     structure.  Appropriate  regulator(s) and  lines  shall  be  provided for
     the type  of cylinder(s)  used.   All analyzers not  utilizing a- refill-
     able  bottle  shall  prominently  display a  label  or  a  sticker  which
     indicates that "AN OPTIONAL  KIT INCLUDING  A  GAS REGULATOR AND HOSE IS
     AVAILABLE   THAT  ALLOWS   THE USE  OF  LARGER  MORE   ECONOMICAL  GAS
     CYLINDERS".   The analyzer manufacturer must  provide such  a kit.  The
     manufacturer's  kit  shall   be  evaluated  during  the  accreditation
     tests.   The  kit  shall  be   included   in  all areas  where  disposable
     bottles of  traceable  span  gas  are  not available from  an accredited
     source.   In  areas  where   disposable  bottles  are   available,  the
     manufacturer's  sales  literature  and  presentations   should at  least
     give  equal  treatment  to  the kit  and  larger bottles  (compared  to the
     disposable  bottle  system).  A  fair   comparison  of  captial  costs,
     operating costs,  bottle  change-out  frequency,  payback time,  etc.  is
     suggested.   If  the  system  includes  a  C02 analyzer,  regulator(s)  and
     lines shall be provided for one  cylinder of C02 span gas.

-------
                                   13
                                                   Spec: E.I. & 3.

E. Section E. Analyzer Performance Specifications (p. 31)

1. Subsection 1. Calibration curve uncertainty, Item a) (p. 31).

     a) Previous Wording; 5% of point (i.e. true value of reading).

     b) New Working; 5% of point  (i.e.  true value of reading)  at and above
     1.2% CO  and  220  ppm HC.  Below  1.2%  CO  and 220  ppm,  the uncertainty
     limits are 0.06% CO and 11 ppm HC.

     c)  Reason for  Change;  The  previous calibration  curve  uncertainty
     specification (5% of point) at  low  CO values was  necessary due to the
     possibility of newly promulgated  Federal  LDT and  HD idle  standards of
     0.47% CO  becoming  performance warranty  standards.  It  is  now reason-
     ably  clear that  5%  of  point uncertainty  at  0.5% CO  is  beyond  the
     capability of optical benches  that  are available  or will  be available
     in the near  future without  major redesign or modifications.  Although
     selected units of  the  current designs can meet  the previous specifi-
     cations, the production  lot  of the equipment is  not  expected to have
     the  consistency  to  meet   the   previous  specifications.    The  new
     specifications are a compromise among accuracy,  cost,  and availabil-
     ity.

2. Subsection 3. Compensation, Item b)i) (p.  31)

     a)  ^Previous  Wording;   The   temperature  compensation  network  shall
     provide accurate results  over the  ambient temperature range specified
     in Section G of  this chapter  as  well as exhaust  gas temperatures up
     to 49°C  (120.2°F).   (Units  with heated  sample  cells are excluded from
     this requirement).

     b) New  Wording;  The analyzer shall provide accurate  results over the
     ambient  temperature  range specified  in  Section G of  this  chapter as
     well  as exhaust  gas temperatures  (at  the  sample cell)  up  to 49°C
     (120.2°F).   (Units  with  heated  sample  cells  are excluded  from this
     requirement).  If the automatic  temperature  compensation  network does
     not  compensate  for the  full  range of the  required temperatures,  the
     analyzer  should  revert  to  a 4  hour  gas  span check  at  temperatures
     outside of the compensated  range.   If the evaluation temperatures for
     Section G  are  outside  the compensated range,  the standard uncertainty
     procedures  (which  include  prior  gas  span  at  the  prevailing ambient
     temperatures)  should be  used at  these   temperatures.   Also,  if  the
     evaluation temperatures  are  outside the  compensated range, additional
     tests should  be  performed to evaluate the  compensation network.  The
     tests should  include checking the  actual  activation  and  deactivation
     temperatures  as  well as  checking  the  analyzer uncertainty  with  the
     compensation  network active  at a  temperature 2°C  above  the  actual
     activation temperature,   and  at a  temperature 2°C  below  the  actual
     deactivation temperature  (e.g. 12.8°C to  35°C  (55-958F)  compensation
     range,  test  temperatures should  be  14.8°C  and  33°C).    The  actual
     activation and  deactivation  temperatures  should  be  +3°C  of  the
     manufacturer's stated temperatures.

-------
                                   14

                                                   Spec:  E.7.  & 8.

     c) Reason  for  Change:  The use  of. a temperature  compensation  network
     is  optional  at  .the  analyzer  manufacturer's  discretion.    Current
     compensation networks are -somewhat limited in  the ability to maintain
     (through electronic  compensation)  an accurate calibration curve  over
     the  full  temperature  range  required  in  Section G.   However,  over
     their limited operating range  (usually 55-95°F),  they  are reported to
     perform adequately.  The advantage of the compensation network is  the
     ability to  shift  from the recommended 4 hour span check  to a weekly
     span check.  The change in the  compensation  specification was  made in
     order to utilize the benefits  of  lower span  gas  consumption  exhibited
     by the compensated analyzers  while maintaining the required accuracy.

3. Subsection 7. Sample Cell Temperature,  Item b)  (p.  32).

     a) Previous  Concept: If  the  sample cell  is  not  heated, the  system
     must provide temperature compensation.

     b) New  Concept: If  the  sample cell  is  not heated,  it   is  preferred
     that the system provide temperature compensation.

     c) Reason  for  Change:  The original  concern  was  about changes  in  the
     sample gas  temperatures  entering the optical  bench.   Current  systems
     tend to  reduce the  sample  gas temperature. to near ambient  tempera-
     tures at (or near)  the  water  trap/filter assembly(s).   Fundamentally,
     this practice  should reduce variability of  the  sample   gas  entering
     the optical bench.   Another  concern was  the potential for the sample
     gas temperature to  be different  from  the  span gas temperature.   Any
     difference  in  temperature would  cause  an error  in  the  span  setting
     that could only  be detected  when  sampling a  vehicle.   Large  span
     errors due to temperature differences should be  able  to  be identified
     by the correlation  test  (to  laboratory analyzers) on  vehicle  exhaust
     (Section  I).    Theoretically,  then,  the  requirement   for temperature
     compensation can be  eliminated, but  to be  on the safe side, we would
     prefer (but  not require)  temperature compensation if  the sample  cell
     is not heated.

     d) New Wording; If  the sample  cell is  not heated,  it  is  preferred
     that the  analysis  system compensates for  temperature  effects  on  the
     gas (sample or span) measurement process  as in Section 3.

4« Subsection 8. Interferences b)  Electronic,  Item v)  (p.  33).

     a) Previous Wording:

     Line voltage and :          1.0% fs L.S.
     Frequency Variation
     (90-130 vA.C.)
     (59-61 hz)

-------
                              15

                                                      Spec: E.8.

b) New Wording:

Line voltage and :        1.0% fs L.S.
Frequency Variation

     1) Normal system operation at 115 vA.C. +^10%

     2) Electronics to be tested between 100-130 vA.C.

     3) Normal system operation at 59-61 hz

c) Reason for  Changei  Most  manufacturers use the  same  source for the
sample  pump.   The  particular  sample  pump  used has  difficulty oper-
ating  at  low  voltages,  and won't  start at  extremely  low voltages.
The  reason  for  the  wide   voltage   specifications   in  the  previous
wording was  that  line voltage  fluctuation  and  location in the power
grid as well as "brown-outs" in urban areas can easily cause the true
line  voltage  to exceed  tradition voltage  specifications  (115 vA.C.
+^10%).  The  concern was about  the  accuracy of  the  analyzer when the
voltage available  for  the  analyzer  was outside  of  the   traditional
val-ues.  Therefore,  to prevent pump  damage, the  evaluation test will
not attempt  to start  the sample pump below 106  vA.C, but  will test
the electronics down to 100 vA.C.

-------
                                   16

                                                   Spec: F.I.3. & 5.

F. Section F. Sample System Performance Specifications (p. 34)

1. Item, Subsection 1. (p. 34)

     a)  Previous  Wording: Maximum sample  cell  mean  pressure differences
     between gas spanning and sampling: 4" H20

     b) New Wording: Maximum  sample  cell  mean pressure differences -between
     gas spanning and sampling:  4"   1^0   or  exhibit  less  than   a  1%
                                 difference in the  mean low-range calibra-
                                 tion  curve  (see  E.I.)  measured  through
                                 the   span   port  compared  to   the   mean
                                 low-range   calibration   curve   measured
                                 through the probe.

     c) Re a s o n for Cha ng e; Allow the manufacturers more flexibility.

2. Item,. Subsection 3. (p. 34.)

     a)  Previous, Wording;  Maximum  sample  cell  mean  pressure  difference
     between normal flow and  low flow indication  :  4" H20

     b) New Wording:  Maximum sample cell mean pressure  difference between
     normal flow and low  flow indication: 4"  H20 or   exhibit  less than  a
                                          1%   difference   in   the   mean
                                          response  on the  low range  to  a
                                          span  gas  measured  through  the
                                          probe  at  normal  flows  compared
                                          with  the  mean  response  on  the
                                          low range  to a span gas measured
                                          through the  probe  at a  flow rate
                                          corresponding  to  the   flow rate
                                          that   activates   the   low   flow
                                          indication.

     c) Reagpn for Change: Allow the manufacturers more flexibility.

3. Subsection .5... System. Leakage.,. Item b) (p. 34).

     a) Previous Wording:  The pressure side of  the  sample  system shall be
     leak  free  as determined by  a  "bubble"  leak-check method  (not  to be
     used in the field).

     b) New Wording;  The pressure side of  the  sample system shall be leak
     free as determined  by a  "bubble"  leak-check method (not to be used in
     the  field).   The sample cell pressure  shall  not exceed  the maximum
     normal operating pressure for this check.

     c)  Reason  for Change^  To protect  the  analyzer  from  possible damage
     due to overpressure.

-------
                                   17
                                                           Spec: F.5. & 6.

4. Subsection 5. System Leakage, Item c) (p. 34).

     a) Previous Wording;  The  vacuum side  leak-check method  shall consist
     of a comparison of  the  span gas response introduced  through  the span
     network to  the  response  of the  same  span gas  introduced  through the
     probe and  sample  line.   In the future with  demonstrated  and  histori-
     cal data,  other leak-check techniques may  be accepted  or equivalent
     to the gas comparison leak-check.

     b) New Wording: The  vacuum side leak-check method  shall  consist of a
     comparison  of  the   span  gas  response  introduced  through   the  span
     network to  the  response  of the  same  span gas  introduced  through the
     probe and  sample  line.   Other  leak-check techniques  may  be  accepted
     as  equivalent  to  the  gas  comparison   leak-check  with  appropriate
     comparative evaluation of  the  fundamental mechanisms  involved,  and a
     demonstration  of  the  technique.   These  other techniques   (such  as
     vacuum level/time decay)  shall use a  1.5% measurement  error trigger
     instead of the 3%  measurement  error  trigger  specified  for  the gas
     comparison leak check.

     c) Reason for Change: Provide the manufacturers more flexibility.

5. Subsection 6. HC Hang-up,  Item a) (p. 35).

     a) Previous Wording; The  HC  analyzer response to room  air sampled
     through the probe and sample line  shall  be less than  20  ppm  C6 prior
     to testing a vehicle or the test is void.

     b) New Wording; The  HC  analyzer response to  room air sampled through
     the probe  and  sample line shall be less  than 20 ppm  C6  (as  measured
     by the analyzer zeroed on  room air) prior to testing  a vehicle or the
     test is void.

     c) Reason  for Change: Most analyzers will  set zero with room air.
     The difference between room air measured from  the  zero  port  and room
     air  measured  from  the   sample  line  will   usually   be  adequate  to
     indicate the amount of hang-up for this type of testing.

-------
                                   18

                                                           Spec: G.I.

G. Section G.. Operating. Environment (p. 36).

1. Subsectionl.b) Ambient Relative Humidity.,. Item, i) .(p.. 36).

     a) Previous.  Wording;  Range of field operation:  0%  to 100% condensing
     (i.e. raining or dense fog).

     b) New Wording:

     Range of field operation:  The  analyzer  shall  be   designed   for  use
                                inside   a   building  or  semi-protective
                                shelter  that is vented  or  open to outside
                                ambient  humidity;    The  analyzer  shall be
                                designed  for  use  in  such  locations  when
                                the   outside   relative   humidity   ranges
                                between  0%  to  100%  condensing (i.e.  rain-
                                ing or dense fog).

     c)  Reason for  Changej   To clarify  intended  instrument  use, .and to
     indicate  that  instrument need not  be  designed for submersible  oper-
     ation.

-------
                                   19
                                                           Spec: H.I.
H. Section H. Fail-Safe Features
1. Subsection 1. Warm-up c) lock-out, Item i) (p. 37).

     a) Previous  Wording:  When  system  power is  turned on.   The  lock-out
     shall stay on  until  the  zero  drift is stabilized.   The manufacturer
     must condition  the lock-out on  analyzer  parameters,  and may  not  use
     clock  time  as  a  sole  criteria  to  determine  warm-up  condition.
     Verification of proper  zero stabilization is determined  by observing
     the  zero  drift over  a  5  minute  period  after  the  lock-out  feature
     deactivates.    The zero  drift   during  this  5  minute  period  may  not
     exceed one-half  of the  zero  drift  specifications  in Section  E.   If
     digital sampling of the  zero level is used, the sample  rate  shall be
     at least 10 hertz.  Analog observation of zero drift is permissible.

     b) New Wording; When  system power  is  turned on.   The  lock-out shall
     stay on  until  the  zero drift   is  stabilized.   The manufacturer  must
     condition the  lock-out  on analyzer parameters,  and may  not use clock
     time as a sole criteria  to  determine warm-up condition.   Verification.
     of proper zero stabilization during  the evaluation test  is  determined
     by observing the zero drift over a 5 minute  period after the  lock-out
     feature deactivates.  The  zero  drift during this  5 minute  period may
     not exceed one-half of the zero drift specifications in Section E.

     c) Reason for  Change; To clarify intent and to  remove  test procedure
     specifications to the test procedure section.

-------
                                       20

IV. Modified Sections in Chapter VIII

    A.  Introduction to Chapter VIII. .(p.. 41).

    1.  Add Sentence: "...  the word  "void"  prominently and/or superimposed over
    the data."   Alternatively,  a different  format for inspection  form versus
    repair form may be used  in  place  of the printer  interlock  and  void stamp,
    provided  that  there  is an  easily   discernable  difference  between  the
    formats and  that  the repair form can  not  be used for  inspection results.
    "All valid test results ..."

    2.  Reason for. Change: Provide more flexibility for the designer.

-------
                                   21

B. Section A. Automatic_Zero/Span checlc (p •

1. Item^ Subsection 6. (p.^41^ 42).

     a) Previous Wording:  The  concentrations of span gas  shall  be entered
     via switches or other convenient means to the following resolution:

          HC = XXXX ppm propane
          CO = X.XX% CO

     The switches  or interlock that  allows  the entering  of the  span gas
     values shall be in an anti-tamper box as described in this chapter.

     b) New Wording:  The  concentrations of  span  gas shall  be  entered via
     switches or other convenient means to the following resolution:

          HC = XXXX ppm propane
          CO = X.XX% CO

     The switches  or interlock that  allows  the entering  of the  span gas
     values shall  be  in  an anti-tamper box  as described  in this chapter.
     In  place  of  locating  the  span gas  concentration  switches in  the
     anti-tamper box,  the  analyzer manufacturer may allow  user  access to
     those  switches  if   the   system  prints   the  gas  concentrations,  as
     determined  by those  switches,   on  each  consumer  inspection ticket.
     The printed value of  the span  gas  concentration need  not  be identi-
     fied as  long  as the  format  allows the auditor to check  the values.
     To  use   this  alternative  (printing   concentration values)   with  the
     optional automatic  data  collection  (ADC) system,   the  microprocessor
     shall  automatically   enter  the  span  gas values  determined by  the
     position of those switches into the ADC storage medium for each test.

     c) Reason  for Change: To  allow the  manufacturers the  flexibility to
     allow  user  access   to   the   span  gas  concentration   switches  while
     maintaining audit capability on the use of those switches.

-------
                                   22

c' Section B. Automatic Leak Check (p. 43).

1. Item, Subsection 3. (p. 43).

     a) .Previous  Wording: Activation of  the  automatic  leak-check system
     shall cause  the  analyzer to  automatically  perform  (or  check)  a span
     sequence, automatically  introduce  span gas  to the  probe,  compare the
     difference between  the  span and  probe readings,  and  make a  pass  or
     fail determination.

     b) New  Wording:  Activation of  the  automatic leak-check  system shall
     cause  the  analyzer  to  automatically  perform (or  check)  a  gas  span
     sequence,  introduce  span  gas  to  the probe,  compare  the  difference
     between the gas  span and  the  probe  readings, and make  a  pass or fail
     determination.  Minimal activity by  the  operator (such as setting the
     probe in a holder,  adjusting  probe flow to  limits  preset  and checked
     by the computer,  capping  the  probe,  etc.  is permitted provided errors
     resulting  from  improper operator  action would  be identified  by the
     computer  and  would  require  corrective  action,  or  improper  operator
     action would tend to cause the system to fail a leak check.

     c) Reason for Change: To clarify original intent.

-------
                                   23

D. Section E. Dual Tailpipes (p. 45)

1. Item, Subsection 1.-4.. (p. 45)

     a) Previous  Concept:  The  dual tailpipe option provided  for  automatic
     averaging of the results from each tailpipe.

     b) New Concept: Delete requirement from Chapter VIII.

     c)  Reaspn  for  Change^  The   number  of  vehicles  in  affected  non-
     attainment  areas  with  true  dual  exhaust  systems is  expected  to  be
     very small.

     d) New Wording: E. [Deleted],

-------
                                   24

E. Section H.. Vehicle Diagnosis, (p. 48)

1. Item, .Subsection 2. (p. 48)

     a)  Pr e vi pus Wo rd i n g:  Activation  of  the  "Vehicle Diagnosis"  switch
     shall allow the analyzer to continuously monitor the vehicle exhaust.

     b)  New  Wor4J-ng:  Activation  of the  "Vehicle  Diagnosis"  switch shall
     allow  the  analyzer  to  continuously  monitor  the  vehicle  exhaust
     regardless  of inspection status  (e.g.  system  needs weekly span check,
     leak check, warm-up condition, etc.).

     c)  Reason  for  Change;  Allows  more  flexible   (but  possibly  less
     accurate) use of the equipment for vehicle repair.

2. Item, Subsection 3. (p. 4.8)

     a)  Previous. Wording: The  printer, or  any automatic  data collection
     system,  shall be prevented  from  operating anytime the analysis system
     is  in a "Vehicle Diagnosis" status.

     b)  New Wording.^ The  printer,  or  any automatic data collection system,
     shall be prevented from  operating  anytime the analysis system is in a
     "Vehicle Diagnosis"  status  unless  one  of  the  options  described in the
     introduction to Chapter VIII is used.
     c)
Reason for Change: Allow printer to be used in vehicle diagnosis.

-------
                                   25

F. Section I. Anti-Tampering (p. 48)

1. Item,. Subsection 2. (p. 48)

     a)  Previous Wording:  All  switches  or  entry  access  for  automotive
     zero/span check  adjustments,  anti-dilution  limits,  span gas  concen-
     tration values, diagnostic  switches, etc.  shall  be  contained  in a box
     or  other  tamper-proof mechanism  with provisions  for an  inspector's
     seal.   A  gummed  label  with  the  inspectors  initials  and date  which
     must be torn to gain  access,  or a braided wire and  crimped  lead seal
     (or similar device) would be sufficient for sealing.

     b) New Wording,; All switches  or entry  access for automotive  zero/span
     check  adjustments,   anti-dilution   limits,   span   gas  concentration
     values, diagnostic  switches,  etc.   shall  be  contained  in  a box  or
     other  tamper-proof  mechanism  with  provisions  for  an  inspector's
     seal.  Span gas  concentration switches may be accessible to  the user
     if the switch values  are printed  for each  test (see option  in Section
     A,  subsection  6.).   A gummed  label  with the  inspectors  initials (or
     authorized  stamp)  and date which must be torn  to  gain  access,  or  a
     braided wire   and  crimped  lead  seal  (or  similar  device)  would  be
     sufficient for sealing.

     c)  ggason  for  Change: To  make  subsection 2.  consistent  with changes
     in Subsections A.6. and 1.3.

       ^ Subsection 3. (p. 48)

     a)  Previous Wording:  The  tamper-proof  system must  allow  convenient
     access by an inspector.

     b)  New Wording:  The  tamper-proof system must  allow convenient access
     by an inspector or authorized service personnel.

     c)  Reason  for  Change:  Allow  service  personnel  to repair or adjust
     analyzers.

-------
                                       26

V.  Modified Sections in Chapter IX

    A. Section B.

    1. Item, Subsection (all) (p. 52)

         a) Previous Wording: B. [Deleted]

         b) New Wording;

         Dual Exhaust Systems

         1.  The  system shall  have  the  capability  to  automatically  calculate
         the average reading for dual tailpipes.

         2. The dual tailpipe  system  shall  use  integrated test values from the
         automatic read system for averaging.

         3. Activation  of  dual tailpipe system shall  allow  two activations of
         the  automatic  read  system  without  activating  the  hang-up  check
         interlock.

         4.  The  dual   tailpipe  system  shall  display  the   average  value  on
         operator command,  and  hold  the value until  reset.   The  average value
         is the value  that  would be forwarded to an  automatic data collection
         system.

         c)  Reason  for Addition; To  provide guidance  for those  programs that
         choose  the option to  have  the  capability   of  testing dual  exhaust
         system  vehicles  and conveniently  enter  that  data  into an  automatic
         data collection system.

-------
                                   27

B. Section C. Anti-Dilution (p. 53)

1. Item, Subsection 6. (p. 53)

     a)  Previous  Wording;  The   C02   analyzer   shall   meet   all  of  the
     analyzer  specifications  in Chapter VII  between  C02 values  of  6% and
     14%.   (CC>2  interference  specification does  not apply).   Specifica-
     tions in Chapter VIII apply to computer analyzers.

     b)  New Wording;  The C02  analyzer  shall  meet  all  of   the  analyzer
     specifications  in Chapter  VII  between CC>2 values  of  5%  and  14%.
     Exceptions  are:  1)   the  C(>2  interference  specification  does  not
     apply, and 2) the uncertainty of  the  calibration curve  shall be +0.9%
     C&2  i° the  range of  5-10%  C02,  and  +0.5%  C02   in   the  range  of
     10-14%  C02-   Specifications  in  Chapter   VIII apply   to   computer
     analyzers.

     c)  Reason for  Change;  The common method  of adding  a  CC-2  analysis
     capability to an HC/CO  measurement system  is  to add a  detector chip
     to  an existing  sample cell.  The  length of  the sample  cell  is very
     critical  in  determining the  accuracy and  discrimination of an NDIR
     analyzer.   Fortunately,  little  compromise  is necessary  in  selecting
     the optimum sample cell  length  for HC and CO.   Thus,  the same  sample
     tube  (cell)  can be  satisfactorily  used for  both HC and  CO measure-
     ment.   Unfortunately,  the optimum   length   for C02  measurement  is
     quite  different  than that for HC  and CO measurement.   Therefore,  if
     we  want the  economy  of  adding just a C02  detector  chip  to  the HC/CO
     sample tube, we must  accept the type  of accuracy that the non-optimum
     C02  path  length gives  us.   The  alternative,  if   more  accuracy  is
     required,  would  be to use  an  additional sample  cell  complete  with IR
     source, optics,  detector,  and electronics;   in short  another complete
     optical bench,  just   for C02  measurement.   The alternative  approach
     while being more accurate, in our judgment is not worth the cost.

-------
                                   28

C. Section D. Loaded Mode Kit (p^ 54)

1. Item, Subsection 4. (p. 54)

     a) Previous Wording: None - new subsection.

     b) Nejw  Wording:  Alternatives to  the  7°C maximum water  trap tempera-
     ture moisture-removal system are:

          i) A  13°C maximum  water trap  temperature  with  corrections  for
          water  content  (electronic  dew  point,  pressure,   etc.)   if  the
          system  can  be  demonstrated  to   provide   equivalent  emission
          results to the base water removal system.

          ii) A..partial  pressure type watertrap  (permature @ etc.)  may be
          substituted for the base water  removal system  provided it can be
          demonstrated to be as efficient in water removal.

          iii)  Any water  removal  system  in  conjunction with a  heated
          sample  cell (with  heated  internal  plumbing)  and  an electronic
          water  correction  system (electronic  dew point,  pressure, .etc.)
          may be used provided  that  the water removal   system:  1)  lowers
          the  sample  gas  temperature  at  least  8°C (14.4°F)  below  the
          sample  cell temperature;  2)  the  internal  plumbing  between  the
          water  removal system and the  sample cell does  not allow the wall
          temperature of  the components  to   drop  below   the  water   removal
          gas temperature;  and  3) the  overall  system provides equivalent
          emission  results to the base water  removal  system.

     c)  Reason  for  Change:  To  allow  the   analyzer  manufacturers  more
     flexibility in designing a loaded mode system.

-------
                                       29

VI . Modified Sections in Chapter XI
         Note:  Throughout  the  test  procedures,  various  test-point
         values  are  required  based  on  the  analyzer  specification
         values.   Revisions  to some  of these  specifications  (found
         in  report  80-5-C)  may   affect   test-point   or  reference
         values.   Therefore,  the  reader is  cautioned to  carefully
         check  the  test procedures  to  assure  that  values  from 80-5-C
         are used for evaluation testing.


    A. Item, Introduction to Chapter XI  (p. 59)

    1. Previous Wording (Last  Paragraph);  The procedures  as written are gener-
    ally independent  of  tolerance  specification  values.   Following  each  test
    procedure,  a  reference  value  corresponding to the  values  given in Chapter
    VII through X will be given in parenthesis.

    2. New Wording:  The  procedures  as written  are  generally  independent  of
    tolerance  specification  values.  Following each  test procedure,  a refer-
    ence value  corresponding to the values  given  in  Chapter VII through X will
    be given in parenthesis.

    It is  important to reemphasize that analyzer manufacturers  may instrument
    candidate analyzers to decrease set-up  time during evaluation (see Chapter
    VII introduction).  An example  of  such  pre-instrumentation is demonstrated
    by the problem of obtaining  an analog chart paper  trace from  a digital
    readout or  CRT display.   Clarification of  the type of pre-instrumentation
    allowed was requested by several manufacturers.   Most digital displays are
    driven  by  BCD  code.   Many  CRT  final stage screen-writing  electronics are
    also driven by  BCD.  Chart  recorders are now available  that  will accept a
    BCD  signal  and print  a stepped  analog trace.   Providing  suitable add-on
    signal  conditioning  equipment  in  order  to  buffer  or  parallel   (with
    parallel  driver if necessary)  the  BCD signal output  to  the  user readout
    device  or CRT screen-writer for the purpose of using  a  BCD chart recorder
    would  be an example of acceptable pre-instrumentation.

    In addition to  the general instructions given above,  results from previous
    testing performed for BAR  80  accreditation can be used in lieu of perform-
    ing  the tests listed  in Table  XI-1.   Results  can only be  used from valid
    tests  performed as  part  of  the  accreditation  procedure  leading  to  full
    accreditation.  The  data from  the  BAR 80 results  is, however,  subject  to
    the 3  year  accreditation period (e.g.  1979 BAR 80 data could be used until
    1982).

-------
                                       30

                                   Table XI-1

                  EPA Test Procedures For Which BAR 80 Results
                               May Be Substituted
EPA Test Procedure No.

       C.2.
       C.3.
       C.4.
       D.6.a.
       E.3.a.
       E.9.

       E.10.
       G.l.ii)
Description of Procedure

Sample Line Crush
Sample Handling Temperature Effect
Filter Check and Hang-up
Sample Line Flexibility
Altitude Compensation
Analyzer Electrical Inference - all except low
  line-voltage testing
Propane to Hexane Conversion Factor
Low Temperature Environment
  - all tests except warm-up and response time
    c) Reason for Change:  To add flexibility and  to  reduce  evaluation testing
    costs.

-------
                                   31
                                                           TP: E.I. & 2.

B. Specific Test Procedure Changes

1. Section E. Subsection 1., Item b) i) (p. 72)

     a)  Previous   Wording:   If   necessary,   follow   the   manufacturer's
     instructions for initial start-up and basic operating adjustments.

     b) New  Wording:  If necessary, follow  the  manufacturer's instructions
     for  initial  start-up  and   basic   operating   adjustments.    If  the
     analyzer  is equipped with  an  automatic read  system,   the  auto-read
     system may be used for this test.

     c) Reason for Change; To add flexibility.

2. Section E. Subsection 3. Item a) v) (p. 77)

     a) Previous Wording: None - adding a new subparagraph.

     b) New Wording: An  alternative  test  procedure  may be used to evaluate
     the   analyzer's   ability   to  be   properly   spanned   at   different
     altitudes.  The  procedure  consists  of  arbitrarily  renaming  the span
     bottle used based on  the expected  density  changes at the various test
     altitudes.  The new values are:

          VH = (BARO/24) (span gas concentration)
          VL = (BARO/31) (span gas concentration)

     Where: BARO is in (in.HgA)

     Use the standard gas  span procedure  to span  to the  Vjj and  to the
     VL values.  See iv) for acceptance criteria.

     c) Reason for Change: To add flexibility and reduce  testing costs.

-------
                                   32
                                                           TP: E.3.

3. Section E. Subsection 3, Item b) i) (p. 78)

     a) Previous  Wording;  Testing concepts:  This  test procedure  is  to be
     performed  in order to  identify  the  performance of  any  pressure  or
     temperature  compensation  systems  under  the  various  environmental
     conditions   that   may  be   encountered   during  vehicle   inspection
     testing.   In  general,  temperature  compensation  will  be  evaluated
     during  the more hostile  environmental  temperature tests .specified in
     Section G.  No other special testing would normally be necessary.

     In  order  to  evaluate  pressure   compensation  systems,   additional
     testing  is  necessary.   If the  analyzer  manufacturer  can  make  a case
     that testing  the  pressure  compensation system in a manner similar to
     the  procedure  specified  in   Chapter   V  Section   E.3.a)  (altitude
     compensation)  will   represent   actual   analys'is  system  operating
     conditions in the  field,  then that procedure (E.3.a) may  be  used for
     check-out.   If a  sufficient  case  cannot  be  made,   and  a  suitable
     alternative  test   procedure  cannot  be  determined,  then  performance
     evaluations  of  the pressure compensation system must  be  carried out
     in  an  altitude  chamber.   The  pressure compenstion  test  shall  be
     conducted  at each  environmental temperature  condition  specified  in
     Chapter  V.   The   tests  shall  be  performed  on  each  range  of  each
     analyzer.

     b)  New  Wording:   Testing  concepts:  This  test  procedure is   to  be
     performed  in order to  identify  the  performance of  any  pressure  or
     temperature  'compensation  systems  under  the  vairous  environmental
     conditions   that   may  be   encountered   during  vehicle   inspection
     testing.   In  general,  temperature  compensation  will  be  evaluated
     during  the more hostile  environmental  temperature tests  specified in
     Section  G.   No other  special  testing  would  normally  be necessary.
     However, if  the temperature  compensation network does not compensate
     for  the  full  range   of   temperature  specified in  Section  G,  the
     activation  and deactivation  temperatures  of   the  system  shall  be
     checked  as well as the calibration curve uncertainty  at  a temperature
     2°C above  the actual activation temperature,  and at  a  temperature 2°C
     below  the  actual   deactivation  temperature.   Checking  of  the activa-
     tion  temperature  etc.  may be  performed  during  the Section  G.  check-
     out.

     In  order  to evaluate  pressure   compensation  systems,   additional
     testing  is  necessary.   If the  analyzer  manufacturer  can  make  a case
     that  testing  the  pressure  compensation system  in a manner similar to
     the  procedure  specified  in   Chapter   V  Section   E.3.a)  (altitude
     compensation) or in any other  manner (such  as component  testing) will
     represent  actual  analysis  system  operating  conditions in  the  field,
     then  that  procedure  (E.3.a.  or  other   procedure  justified  by  the
     manufacturer) may  be used  for  check-out.   If a sufficient  case cannot
     be  made,  and  a   suitable   alternative  test  procedure   cannot  be
     determined,  then performance evaluations of the  pressure compensation

-------
                                   33
                                                           TP: F.5. & G.I.

     system must  be  carried out  in  an altitude  chamber.    The  pressure
     compenstion test shall  be conducted at  each environmental temperature
     condition  specified  in Chapter  V.   The  tests  shall be  performed on
     each range of each analyzer.

     c.  Reason for  Change: To  be  consistent  with previous  temperature
     compensation changes and to allow more flexibility  in  the  testing of
     pressure compensation networks.

4. Section F. Subsection 5., Item b) xiv) (p. 104)

     a) Previous Wording: None - adding a new subparagraph.

     b) New Wording;  Repeat steps  i) through xiii)  with the  tee fitting
     between the filter assembly and the leak check sensor.

     c)  Reason  for   Change;  To  compensate  for  and  to test  adequately
     alternative  leak-check systems  that  are  allowed  due  to changes in
     Chapter VII.F.5.

5. Section G. Subsection 1., Item b) (p. 105)

     a) Previous Wording;

          Test Conditions

          i)  105°F  (^5%°F)  with a  relative  humidity  between  80  and 85
          percent (non-condensing).

          ii)  40°F  0*5%°F)  with a  relative  humidity  between  75  and 80
          percent with a 10 mph wind.

          iii)  35°F   (HH5%°F) with  a  relative  humidity between  10  and 20
          percent.

     b) New Wording:

          i)  105°F  (jf5%°F)  with a  relative  humidity  ^between  80  and 85
          percent (non-condensing).

          ii)  40°F  0+5%°F)  with a  relative  humidity  between  75  and 80
          percent with a 10 mph wind.

     c)  Reason for  Change: Test  conditions  ii) and  iii)  were  . somewhat
     similar,  therefore  to   reduce evaluation  testing time  and  cost,  test
     condition iii) was dropped.

-------