EPA-AA-IMS-/81-16
              Technical Report
       DRAFT EPA RECOMMENDED  PRACTICE.




       FOR NAMING  I/M CALIBRATION GAS
               September, 1981
             Thomas L. Darlington




       Inspection and Maintenance  Staff




    Emission Control Technology Division




Office of Mobile Source Air Pollution Control




     Office of Air,  Noise, and Radiation




    U.S. Environmental Protection Agency




         Ann Arbor, Michigan  48105

-------
                                   Abstract
This  report  explains  how calibration  gases will  be used  in  Inspection and




Maintenance  (I/M)  programs,  and identifies  the problems  states may  have in




obtaining   accurate    gases.    A   Recommended-  Analysis   Practice   for  gas




manufacturers  to  use when naming  I/M calibration gases  is  presented.   States




are encouraged to procure gases  named according to  this Recommended  Practice




for their own  use, and to require  licensed inspection stations to procure  them




to  ensure  that they are  obtaining accurate  calibration gases which  meet the




terms  of  the  Emission  Performance  Warranty   [207(b)],  and  to  improve  the




general quality of their  I/M programs.

-------
                               Table of Concents
1.0   BACKGROUND                                                         4
1.1   EPA Regulations Concerning Calibration Gas Accuracy                5
1.2   Practical Problems With Warranty Accuracy Requirements             6
1.3   EPA Efforts to Help States Meet Warranty Requirements              7
      for Gas Accuracy

2.0   INTRODUCTION                                                       8
2.1   I/H Calibration Gas Users                                          8
2.1.1 Centralized Programs                                               8
2.1.2 Decentralized Programs                                             8
2.2   Components of I/M Calibration Gases                               10
2.2.1 Carbon Monoxide and Propane                                       10
2.2.2 Carbon Dioxide                                                    12
2.2.3 Hexane                                                            15
2.2.4 Diluents (Balance Gases)                                          17
2.3   Concentrations of I/M Calibration Gases                           18
2.3.1 Emission Performance Warranty Requirements                        18
2.3.2 State Requirements                                                19
2.4   NBS and Other Standards                                           20

3.0   TRACEABILITY                                                      24
3.1   Establishing Traceability: Overview of Procedures                 24
3.2   Analysis of Pure Components                                       26
3.3   Cylinders                                                         27
3.4   Instrument Preparation and Calibration                            28
3.4.1 Definition of Linearity                                           28
3.4.2 C02 Interference Check                                            30
3.4.3 Calibration Curve for Linear Instruments                          30
3.4.4 Calibration Curve for Non-Linear Instruments                      32
3.5   Analysis of I/M Calibration Gas Cylinders                         34
3.5.1 Re-Usable Cylinders                                               34
3.5.2 Disposable Cylinders                                              35
3.6   Calculating I/M Calibration Gas Concentrations                    38
3.6.1 Equations Used'                                                    38
3.6.2 Concentration Determination                                       38
3.7   Cylinder Labeling and Documentation                               39
3.3   Audits of I/M Calibration Gases                                   40
                                  Appendices

Appendix 1    Emission Performance Warranty Regulations
              Pertaining to Calibration Gases                           41
Appendix 2    Listing of Analysis Requirements for
              Recommended Analysis Practice                             42
•Appendix 3 .   Accuracy Discussion                                       55

Appendix 4    Discussion of Statistical Sampling and Naming

-------
1.0 BACKGROUND








An Inspection and Maintenance  (I/M)  program is a  state  or locally run program



in which  registered  vehicles  are  required  to  obtain  and  pass a  tailpipe



emission  inspection once  per  year.   Vehicles  that  have  tailpipe  emissions



greater  than  state  or  locally  established emission standards  are required to
                                   i


obtain maintenance  to pass  that standard.  The emission  test  (called a "short



test") can be conducted  by the  state (or  locality)  or  a contractor  to the



state, and this is  referred to as a  "centralized  program".   The emission test



can  also be  conducted  by  private garages  which  are  licensed by  the  state.



This  latter  system is  called  a  "decentralized"  program.   Twenty-nine  states



are expected  to have operating  programs  by Janurary 1,  1983,  however,  many of



these programs will commence before  that date.   At this  time, 11  states are



expected  to  have   decentralized   programs,  16  states  are expected  to  have



centralized programs, and two states are undecided.








The  primary  inspection  and  diagnostic   tool of   these  I/M  programs  is  a



Non-Dispersive Infrared (NDIS)  analyzer,  which is  capable  of  determining the



concentrations of hydrocarbons  (as hexane) carbon monoxide, and  in some cases



carbon  dioxide  in  raw  vehicle  exhaust.   These  analyzers,  however,  need



periodic maintenance to keep their accuracy, as operator misuse,  or changes in



pressure,  temperature,   and  other   operating   variables  can   render  them
                                    c*


inaccurate.








Many analyzer manufacturers  recommend  a periodic  check of their analyzers with



an accurate calibration gas.  State or local I/M  program regulations  will also

-------
require  a  periodic  check.   This  check  is   performed  by  flowing  a  known

concentration of calibration  gas  into  the analyzer and determining whether the

analyzer  is  reading  this calibration  gas correctly.   If  it  is not,  then a

simple  adjustment  (i.e.,  a  calibration)  of  the  analyzer  can  usually  be

performed which  will result  in the analyzer  being  accurate  again.   However,

the  accuracy of  the  calibration  gas  used  is  also  important  in  determining

analyzer  accuracy  with this  maintenance  check.  If  the  labeled concentration

of  the  calibration gas  is significantly  different than the  concentration in

the  cylinder,  the   analyzer  could  become  significantly  misadjusted.   An

operator  who  had  performed the  calibration would  be unaware  that  he/she  had

actually  misadjusted   the analyzer,  since  he/she trusted  the  label on  the

calibration cylinder  to be correct.



1.1 SPA Regulations Concerning Calibration Gas Accuracy



The  EPA  recently  promulgated  Emission System  Performance Warranty  Regulations

for  1981  and  later vehicles which entitle a vehicle  owner  to emission-related

repairs  at   the  manufacturer's  expense  if,  among other  things, the  vehicle

fails an  "approved" emission  short test.   One of  those conditions is  that  the

analyzer  used  to  conduct the emission short  test  must be  checked with  gases*

"traceable to NBS  standards +2Z" at  least once per week (Section 85.2217,  page
*  Note:   There  are  no  C02 emission  standards  in  the  Emission  Performance
Warranty, consequently,  the Warranty  accuracy  specification of ^22  traceable
to  NBS  does not  apply to  C02«   Also, since  there are  no  NBS standards  for
hexane,  the Warranty  accuracy.specification of traceable ^22 to NBS  cannot be
met on an analyzer calibrated with hexane only.

-------
34808.  The  relevant portipn  of  this page  is  provided in Appendix  1  to this




document).   There are  other  requirements  for  these  gases  in  the  Warranty;




however,  these  are  beyond the scope  of  this section and will  be addressed in




Section 2.3.1.  The  EPA  thinks that many states implementing I/M programs will




want to make Warranty  protection  available to consumers participating in their




I/M  programs.   These  states will  try  to  make sure  that the  gases used  to




periodically calibrate analyzers used in  the  I/M programs meet  the Warranty




requirement of having an  accuracy of ^22 to NBS standards.  However,  there are




some  practical  problems  associated with  this  accuracy requirement  which  are




discussed in the following paragraphs.








1.2 Practical Problems With Warranty Accuracy Requirement








States might  think that  they  have  only  to order calibration gases  for their




analyzers that have been  certified  by the manufacturer  to  be  "traceable  to  NBS




standards  ±21"  to  be  ensured of  meeting the  accuracy   requirement  in  the




Warranty.  However,  the most conscientious gas  manufacturers  are very hesitant




to label  cylinders of  calibration  gas that  they have analyzed  with  the  phrase



"^22 to NBS standards", as they point  out that  there is currently no clear-cut




definition  of  "traceable"  to  NBS  stanards.  They  explain  that  traceability




implies a system  of analytical procedures  and documentation  of data that  is




used  to express  the  concentration  of  a  particular gas  relative  to  a standard




gas.   These analytical   procedures and  methods  of handling   data  may vary




significantly between manufacturers,  causing differences in the concentrations




of  identically  labeled calibration gases made by different  manufacturers  or




even  by  the  same manufacturers.   I/M  calibration gas  users   would find  it

-------
difficult  to  detect  these  differences and  identify which  gas  cylinders were




correctly,  labeled,  with  the  possible exception  of  the  I/M managers  if they




maintain a sophisticated gas analysis  program of  their own.








1.3 EPA Efforts to Help States Meet Warranty Requirement  for Gas Accuracy








In order to improve the general  quality  of I/M  programs,  and to help states  in




meeting  the  accuracy requirement  of  the Warranty with  respect  to calibration




gas used  to  check emission analyzers,  the EPA has  published  this Recommended




Practice which  states can  require gas manufacturers  to  follow in naming and




labeling  calibration  gas  for  the  I/M  programs in  those  states.    A   state



requirement would be  imposed directly  by  the  state in its own purchases  of gas




and/or  indirectly by  establishing rules  and  regulations which  require   other




I/M gas  users (contractors or  licensed  inspection  stations)  to  buy  only gas




that  the  manufacturer  certifies was named  and  labeled  according   to this




Recommended Practice.








A  special effort has  been made  to  make these procedures  workable from  a




scientific gas  manufacturer's viewpoint.   A  meeting was held between the gas




manufacturing industry,  the National  Bureau of Standards and  the  EPA  to  agree




on  the original  framework of  the procedures.   Manufacturers have  also been




allowed  to comment  on the draft  forms  of the  Recommended Practice:  many   of




their  comments have  been incorporated into the procedures.  The  end  result  is




a  Recommended Practice  which  should  contribute  significantly to  overall I/M




program accuracy  and  fairness.

-------
2.0 INTRODUCTION









2.1 I/M Calibration Gas Users









2.1.1 Centralized Programs









In a centralized I/M program where  the  official I/M short test is conducted in




state  or contractor  operated  centralized  test  facilities  only,,  states  are




likely to require that  only  those  analyzers in the centralized test facilities




be  checked  at  least  once per week  with an  accurate calibration  gas.   Most




states operating centralized  programs will  not require  garage owners  who have




purchased analyzers  to  help them  in the  repairs  of vehicles, failed  from the




I/M  programs  to  perform  periodic  calibration checks  on  their instruments.




Therefore,  analyzer  operators  in  the  centralized  facilities  will  be  the




primary  users of  calibration gas  in a  centralized  program.   The- exception to




this situation  is where  a  state running a centralized program licenses certain




garages  for the  reinspection  function  (New  Jersey, for  example).   In this




case, the state will  likely  require reinspection garages to perform a periodic




calibration check  of  the  garage  analyzer  with  an  accurate  calibration gas,




also.









2.1.2 Decentralized Programs









In  decentralized programs, there are likely to be  two different categories of




I/M calibration gas users.  The first are the  garages licensed by the  state to




perform  emission inspections.   States are likely to require that these garages

-------
perform a periodic  check of  the  garage analyzer with  an accurate calibration

gas.    Also,   decentralized   programs   will  have   official   auditors   who

periodically  visit each  facility  to determine  (among  other things)  if the

emission analyzers  are  accurate.   Part of  the  auditor's job  will  be to  check

the  facility's  analyzer with  state  audit  gas,  to  provide  a  double   check

against the facility's gas.



Many  states  are  requiring  their  auditors  to  use   calibration  gases  with

accuracies of ^12 to NBS  standards  rather than the  accuracy requirement of ±22

which is specified  for weekly calibrations  in the Warranty.  Since some states

have proposed  regulations which would require  auditors to  use calibration gas

which is  traceable HZ,  included  in this  Recommended  Practice are procedures

which will  allow gas  manufacturers to name certain audit  gases  to within ±1Z

of NBS standards.*



Decentralized  programs   are   expected  to   use  more   calibration  gas  than

centralized programs.   This  is because decentralized programs  consist  of many

small neighborhood  garages performing a  few emission inspections  each,  where a

centralized program consists  of  a  few high-throughput inspection  facilities.

Consequently,  there  are  more  analyzers  in  a  decentralized  program,  each

requiring  the  same  periodic  calibration  checks  as  the  fewer  number  of

analyzers used in a centralized program.'          .
*  The  one exception  is an  I/M calibration cylinder  which contains  both  C02
and  CO.   The  best accuracy 'specification  on this cylinder named  according to
the Recommended Practice is  H.5Z.   This  issue  is discussed further in Section
2.2.2 and Appendix 3.

-------
                                      10

2.2 Components of I/M Calibration Gases



A state has several options  in determining how many and what components are in

the calibration  cylinders which are  bought for calibrating  analyzers  in that

state.  The Recommended Practice generally  does  not limit the  scope of these

options: within  certain limits as discussed in this section,  there  should be

no  difference in  the  quality or  accuracy  of gases  in each  option  if  the

components  are  named-  in  accordance  with  the procedures  described  in  this

Recommended Practice.



2.2.1 Carbon Monoxide and Propane



States  must   first  examine  what  components  are  needed   in  a  calibration

cylinder(s)  to  ensure  that  inspected  vehicles   are  accurately  passed  and

failed.  Some  I/M  programs  are being  implemented with a  carbon monoxide  (CO)

cutpoint only.*  The  state  may feel that  analyzers therefore only have  to be

accurate at reading a vehicle's CO  emissions, and determine that a calibration

cylinder of CO is  sufficient  to  ensure this  accuracy.   This determination is

consistent  with  the  Emission  Performance  Warranty.    However,   the   EPA

recommends that  a  state with a CO-only program require inspectors to calibrate

their  analyzers with  propane  as  well  as  CO,  since a mechanic  performing

repairs on a vehicle can use a  vehicle's HC  emissions to diagnose problems
*  Some  urban areas that  are  in non-attainment  for  CO but not  for  ozone will
implement  an I/M  program which uses  CO cutpoints  only.   The state  of  North
Carolina is  operating  an  I/M  program which is  in  the  voluntary inspection and
maintenance  phase  of implementation which has cutpoints for CO only.

-------
                                      11

that may  be  causing a significant  loss  in fuel economy  for  that vehicle, and

the  costs of  calibrating  an analyzer  with  propane  as  well as  CO  are not

significantly greater than the costs of calibrating an analyzer with CO only*



Most I/M  programs will be implemented with CO  and  EC  cutpoints.   These states

therefore are likely to require  all the  analyzers  used for inspections in that

state  to  be  calibrated  with  both  CO  and  propane**    The  most  convenient

cylinder  for them to order will  be  a three-blend cylinder of CO and propane in

nitrogen.  Some  states may  wish  to order  separate  cylinders  of CO in nitrogen

and propane  in nitrogen on  the  theory that there  is  "interference" between CO

and  propane  in-  nitrogen  such  that both  components  appear  to  have  slightly

higher  concentrations than  they  would  if  they were  in  separate cylinders.

[However,  the  National Bureau  of  Standards   (NBS)  has   studied  the  level  of

interference   between  CO   and   propane  in   typical    I/M   calibration   gas

concentrations,  and  determined  that  the   interference   effects   are  very

slight.**]   Consequently,  little accuracy is  gained by  using  two binary, blend

calibration gases (i.e.,  CO  in nitrogen  and  propane in  nitrogen)  than by using

a single  tri-blend calibration gas  (i.e.,  CO and propane  in nitrogen).
* Hexane  might  be used sometimes  instead  of propane.  Hexane  is  discussed in
Section 2.2.3.
**  [Reference to not-yet-conducted NBS study.]

-------
                                        12

  2.2.2 Carbon Dioxide



  Some I/M  programs  will have  carbon dioxide  (CO-)  outpoints  for  vehicles  as

  well as HC  and  CO  cutpoints.   Checking for the  amount  of CO, in  a vehicle's

  exhaust helps the state to  determine whether there are  significant leaks  in a
                                                                      \
  vehicle's  exhaust system which are causing dilution of exhaust gases.



  The   Recommended  Practice   contains    procedures   which  will   allow   gas

  manufacturers  to  incorporate  CO.  into  I/M  calibration  gas  mixtures  and

  demonstrate  traceability  j^2Z  to  NBS   of   all   components   used.   However,

  traceability better than  ±1.5Z cannot  be attained for  these  mixtures  because

  of interference  effects  between  CO   and CO.   These and  other  considerations

  are  discussed in the following paragraphs.



  Inspection  analyzers  use  infrared   absorption   principles  to   detect   the

  concentrations  of   propane,  CO,  and  CO-  in  vehicle   exhaust.    These  gases

  absorb  infrared  radiation at  different wavelengths,  consequently  the  analyzer

  can,  for  the  most  part,  differentiate  completely between each gas.   However,

  the  wavelengths  of  absorption  of CO  and CO-  are very close  to  each  other.

  This has  the  effect  of making  some  infrared  analyzers  slightly  inaccurate  at

  reading CO  in  the presence   of  CO ,   because they  actually  count  a  small

  portion of  the  CO   in   the   analysis  of  CO  (but  not vice  versa).*   This

  phenomenon is known as interference.
,  * The ability  of  infrared analyzers to read  these  components discreetly  is  a
  function of the quality of filters used in the analyzers.

-------
                                      13




Laboratory-type infrared  analyzers  which gas manufacturers will  likely use to




name  CO  in  an  I/M  calibration  cylinder  containing  CO.  and  propane  are




capable of  keeping interference  to  minimum levels.   The Recommended  Practice




requires  gas  manufacturers  to keep  the level  of CO.  interference in  CO to




less than .31 of  the CO concentration when naming  CO in an  I/M calibration




cylinder  containing   CO^*    Also,   the  Recommended   Practice  requires  gas



manufacturers  to  use  tighter  tolerances in  naming I/M  calibration cylinders




containing  CO..   The  small amount-  of  additional  interference  error  present




when  CO  is  used requires  these tighter  tolerances  so that  all  components




meet the  ±2£  to  NBS accuracy  requirement.   However,  the tolerances  within the




Recommended   Practice  have   not  been  tightened  to   such   a   degree   that




traceability  _^1I  for  a  mixture containing  CO.  and  CO could  be   assured.




Consequently, M..5Z  is the  best specification for this mixture attainable  with




this  specification.    A  more  complete  discussion  of  how  these   accuracy




specifications are arrived  at  is  presented  in Appendix  3.

-------
                                      14

Regardless   of  how   accurately   the   calibration   gases   are   named   by

laboratory-type  infrared  analyzers  used  by  gas  manufacturers,  inspection

analyzers are  still likely to  exhibit  small interference  effects.*  A recent

study  by  EPA  determined  the  CO   interference  levels  in  several  common

inspection analyzers.   The results of the study are presented in Table 1.**
* Another  question raised  by  interference effects  of C02 on  CO is:  What  is
the effect  of  calibrating an  analyzer  with /a calibration  cylinder  containing
CO  and  C02 on  a  vehicle's  apparent   CO/- 'amiss ions?   If  an  analyzer's  CO
response is  calibrated with a  CO cylinder containing  C02»  the  analyzer will
be set at  slightly lower than  the  actual  CO concentration in  the calibration
cylinder because  the  analyzer will  be responding  to  a small  amount  of  CO2
also.   Since the  analyzer's CO  response will be  set  slightly  lower,  it will
read  a  vehicle's  CO  emissions  at  a slightly / lower  level,  -thereby making  a
given short  test slightly  less stringent  than  it  would  have  been had a  CO
cylinder that did not. contain C02 been used for the calibration.

** "Operational Evaluation  of  Vehicle Exhaust Emission  Inspection Analyzers",
Volume 1, Contract # 63-03-2747, EPA 460/3-80-019.

-------
                                      15

                                    Table 1

                            Combined Z Full Scale*
                           . EC and CO Responses of
                             Analyzers to  15Z C02

                   Analyzer                 Z Combined F.S. Response

                       A                              O.OZ
                       B                              1.74Z
                       C                              2.04Z
* The combined interference  levels  reflect the percentage full scale responses
of  the  EC and  CO  meters  to CO2 gas.   The  percentages  were  added to  get  a
"combined" response.
Based on  the  data in Table  1,  it is probably  better for an  I/M inspector to

check the CO response of an  analyzer  without  using a CO cylinder that contains

CO..   If  a  state  determines  that  CO   must   also  be  used,   then  the  CO

should be  in a separate  cylinder from the  CO   (i.e.,  one  cylinder of  CO and

propane  in  nitrogen,  the  other  cylinder CO.  in  nitrogen).    If the  state

determines  that  the CO.  must be  in  the  same  cylinder as  the   CO,  the state

should  select  the  lowest   CO.   concentration  acceptable   to   maintain, the

analyzer's  CO.  accuracy.  This  will  minimize  the  interference  effects  of

CO. on CO.



2.2.3 Besane



Idle   emission   inspection   analyzers,   when   determining   the  amount   of

hydrocarbons  (EC)  in  vehicle  exhaust,  are  really measuring  the amount  of

hexane  (actually  n-hexane)   in  vehicle  exhaust,  plus  some  fraction  of  the

-------
                                      16




remaining  hydrocarbons  which  appear   to   be   partially  hexane  because  of




interference effects.  The assumption inherent  in the measurement technique is




that the quantity of  actual  hexane plus other hydrocarbons which appear to be




partially hexane  due  to interference  is  roughly proportional  to  the total




quantity of all of the various types of hydrocarbons.








Because  analyzer  readouts  are  based  on  apparent  hexane  concentrations  in




vehicle exhaust, hexane  is in theory  the best gas with which  to  calibrate the




HC portion  of the  analyzer.   However,   there  are several  problems  associated




with hexane  which make  it  impractical  as  an inspection  analyzer calibration




gas,  except under carefully controlled conditions.








First,  there  are  no NBS  standards for  hexane.   This means that  the accuracy




specification  of   "traceable ±ZZ  to  NBS"  is   by definition  impossible  for




hexane.  However,  that does not necessarily mean that one cannot  obtain"highly




accurate hexane calibration gases.  There are  scientific  gas  manufacturers who




have the resources to blend and name hexane accurately.








Second, hexane is a "sticky" molecule,  and  therefore has a tendency to cling




to the  sidewalls  and interior  fittings  of a cylinder  if the cylinder  is  not



periodically  agitated and/or  heated.    If  sticking  occurs  in a cylinder  of



hexane, the concentration of hexane in  the  cylinder  changes as  the cylinder is




used,  since   the  diluent (balance gas) will  flow  from  the  cylinder  first,




leaving  the  hexane  behind.   Agitation  of  the  cylinder  by  rolling it,.,  and




heating of the cylinder can significantly reduce this problem.

-------
                                      17




Lastly, hexane is much more expensive than  propane  to use as a calibration gas




because of  the  problems in handling and blending.it.   Nonetheless,  it still




has potential uses in an I/M  program.   For  example, because propane is usually




used  instead of  hexane  as  a calibration gas, analyzer manufacturers  must




develop correction factors for  their analyzers  to convert a propane reading on




the analyzer to a  hexane  reading.  This correction factor can change slightly




if the  analyzer  is roughly handled.  A hexane calibration  cylinder  is  needed




to check  or  re-establish  this  correction  factor  periodically.    Arizona  and




California   I/M  programs   are  currently  using   hexane   to   check  analyzer




correction factors.








2.2.4 Diluents (Balance Gases)








Throughout this report we have  assumed  that calibration gases  would be blended




with nitrogen as a diluent or balance gas.  Nitrogen is the easiest gas  to use




as a  diluent from a production  standpoint since  it  is readily  available  in




very pure forms.








Air could  also be used as a diluent,  however,  gas manufacturers  cannot  use




room air or  ambient  air because of the  impurities  present  in  these sources  of




air which could  cause errors  in  the blending  and  naming  processes.   In order



to use  air  as a diluent,  gas  manufacturers must buy pure  nitrogen and  oxygen




and  blend  it  together prior  to  filling   I/M  calibration  cylinders.    This




process adds to  the production  costs  for  I/M calibration gases,  therefore,




most states will order I/M calibration gases which use nitrogen as a diluent.

-------
                                      18





2.3 Concentrations of I/M Calibration Gases









2.3.1 Emission Performance Warranty Requirements









In  addition to  the  accuracy requirement  for gases  used  to  check  emission




analyzers  (i.e. ,  ±2Z  to  NBS  standards)  the  Warranty  also  requires  the




calibration gas  to be  vithin certain  concentration ranges.   The  limitations




are  that  the  calibration gas oust  have  concentrations of  between  -50Z  and




+100%  of  the  state emission  standards if  the  state  adopts  the  numerically




minimum emission  standards  specified in  the Federal Warranty  Regulations,  or




have concentrations between  these numerically minimum standards  and  the state




selected  emission  standards   if  the  state  selected emission  standards  are




numerically higher  than the  minimum emission  standards.   The minimum emission




standards specified in  the warranty  are,  depending on the short  test procedure




used, either l.OZ  CO  and 200 ppm hexane  HC or 1.22 CO  and  220 ppm hexane for




HC.  If a state  selects emission standards of l.OZ  CO and  200 ppm hexane,  the




concentration  of calibration  gas  used  to  periodically check the  analyzers must




be  between  0.5Z  and 2.0Z for  CO,  and  between  100  ppm  and 400  ppm hexane  (or




200 ppm and 800  ppm propane)  for the emission short test  to  meet the  terms of




the  Emission  Performance Warranty.  Alternately,  if a  state  selects  emission




standards of 4.0Z CO and  400  ppm  HC  (the  highest  standards any state is likely




to  adopt  for  1981  and  later vehicles),  the concentration of  calibration  gas




used to periodically  check  the analyzer must be between  l.OZ  and 4.0Z for CO,




and  between  200 ppm and  400  ppm  hexane for HC  (or 400  and  800 ppm propane).




See  Table 2.   The  lower limits of l.OZ CO and  200  ppm hexarie  could be 1.2Z CO




and  220 ppm hexane  if a different test  procedure is  used.

-------
State Adopts

Warranty Emission
Standards of

Warranty Emission
Standards of

State Emission
Standards of
                                      19

                                    Table 2

                    Example of Calibration Gas Requirements
                         for  207(b) Warranty  Coverage
  Emission
  Standard

200 ppm £6. HC
l.OZ       CO

220 ppm C6. HC
1.2Z       CO

400 ppm C£ HC
4.0Z       CO
Calibration Gas Concentration.*

   200 ppm C3- 800 ppm C3_
   .52 CO    - 2.0Z CO

   220 ppm C3- 880 ppm £3
   .62 CO    - 2.4Z CO

   200 ppm C3- 800 ppm C3_
   l.OZ CO   - 4.0Z CO
  Note: Standard.'is expressed  in ppm hexane (C6).  Span gases are expressed  as
        ppm propane (C3) assuming an  approximate  correction  factor of  0.5.
2.3.2 State/Requirements
In  addition   to  meeting  the  Warranty   requirements   for  propane  and   CO

calibration   gas  concentrations,   some   states   may   wish   to   use  other

concentrations  of propane  'and  CO  in performing  multipoint  calibrations  on

analyzers  used in the  I/M programs.   Also,  some  states  may wish  to  use  CO

in multipoint  calibrations  also.   In  these cases  the  state would need several

cylinders   of   propane,   CO   and/or   CO,    calibration  gas   at   varying

concentrations,  all with accuracies  of _^2Z to  NBS  standards for best  results.

For  these reasons, we  have designed  the Recommended Practice  in  such  a  way

that  a  gas  manufacturer can  demonstrate  traceability  for  any ordered   gas

concentration  which falls between the  range of NBS standard reference  material

concentrations  discussed in  Section 2.4  (see  Table  3).   I/M  calibration  gas

-------
                                      20





concentrations ordered  either above the  highest  standard or  below the lowest




standard listed in  Table  3 are not  assured  traceability ^22  to  NBS standards




within the scope of this Recommended Practice*









2.4 NBS and Other Standards









The National  Bureau of  Standards tries  to  make  available  standard gases  in




commonly requested  concentrations.   Up  until  this  year,  the  only NBS standards




available   were    called   Standard    Reference    Materials   (SRM's).    The




concentrations of  various  NBS CO,  propane,  and  CO-  SRM's  which  are  in the




range  of  interest  as far  as   I/M calibration gas  is  concerned are  listed  in




Table 3.  SRM's have an analytical uncertainty of  ^1%  to the true value of the




concentration in the cylinder.

-------
                                      21

                                    Table 3

                NBS Propane, CO, and C02 Standards in Nitrogen
                          in the Concentration Range
                       of Typical I/M Calibration Gases
                  Propane
               NBS Standard

                   2643
                   2644
                   2645
                   2646
                   2647
                   2648
   Propane
Concentrations

    100 ppm
    250 ppm
    500 ppm
   1000 ppm
   2500 ppm
   5000 ppm
                  Carbon
                 Monoxide
               NBS Standard

                   2638
                   2639
                   2640
                   2641
                   2642
   Carbon
  Monoxide
Concentration

 5000 ppm C.5Z)
    l.OZ
    2.0Z
    4.0Z
    8.0Z
                  Carbon
                  Dioxide*
               NBS Standard

                  2620a
                  2622a
                  2624a
                  2626a
                  1674b
                  167 5b
   Carbon
  Dioxide
Concentration

    l.OZ
    2.0Z
    3.0Z

    7.0Z
   14. OZ
* Other  standards are available  for C02-  A  list  of available  standards  may
be obtained  by writing  the  Office  of Standard  Reference  Materials,  Chemistry
Building, Room B311, National Bureau of Standards, Washington, B.C. 20234.

-------
                                      22



Historically, SRM's have  been the only  NBS standards.  The  EPA  has therefore


required SRM's as  the  standard  to which calibration gases  used  in many mobile


and stationary source  programs  had to be traced.  This  created  a large demand


for SBM's  since the  tracing process  consume the  SSM's.   Also,  the analysis
                                                        \

procedures  for  SBM's  are very  detailed and  lengthy, so  SRM's  have  been in


short supply.   In  an  attempt to  alleviate  these problems  associated with the


lack  of supply  of NBS   standards,  the EPA  Emission  Monitoring  and  Support


Laboratory  at Research Triangle  Park  (EMSL-RTP)  and  NBS  developed procedures


that would  allow gas manufacturers  to make a new  kind of standard,  called  a


Certified Reference Material (CRM), which  if made  properly  has  an analytical


uncertainty  of  about  ±1.3Z  to  true  value.   These  CRM'3  can  be made  in  batch


quantities,  but  a  limitation of  these CSM's  is  that  they can be made  only at


SRM concentrations.






The  analytical  procedures  described in  this report  rely on  there being  an


adequate  supply of NBS   standards,  either  SRM's or   CRM's.   The  CRM  program


should ensure that availability.*





Because  of  the  difficulty  in  obtaining  NBS   standards,   scientific   gas


manufacturers have,  over  the years,  developed  their  own in-house  standards


which  they  have  used in the  analysis  of  various  scientific  gases.   These


in-house standards, called primary standards, are for the most part very
* At the time of publication  of  this  report,  one  gas  manufacturing company was

offering CSM's for sale.

-------
                                      23




accurately  named.   However,  there  are  no uniform  procedures  throughout  the




scientific gas industry for naming  these standards.   Therefore,  differences in




accuracies  of primary  standards between manufacturers  probably exists.   We




mention these standards in  this  section  because  they are used by manufacturers




along  with NBS  standards  (if  a  gas  divider  is  not  used)   to  demonstrate




traceability  within  the   context   of  this  Recommmended  Practice.   The  UBS




standards  serve  the  function of  providing a  check  on the  accuracy of  the




in-house  primary  standards*   The  primary  standards  serve   the  function  of




providing more data points  by which  to construct  an analyzer  calibration curve




than would be obtained if NBS standards were used alone.

-------
                                      24




3.0 TRACEABILIT7









People familiar with  the  blending and naming (the  affixing  of a concentration




label to a cylinder) of scientific  gases  use the word traceability to describe




an analytical link  between  the concentration of  two  separate  cylinders  of the




same kind  (not  necessarily the  same  concentration) of  gas,  where one  of the




cylinders  is  usually  some  kind  of standard gas.   The  standards most  often




referred to in  any  traceability definition  are  NBS standards  (these  standards




were discussed  in section 2.4).  For  the purpose  of  this  report,  "traceable"




will mean the use of  the  standards  and analytical procedures discussed in this




report.  These  procedures,  if used properly,  should yield  an I/M calibration




gas  which  is  accurately named  relative  to NBS  standards  (either  SRM's  or




CRM's) within ^22, _+1.5Z, or  ±1%,  depending on  the criteria selected  with the



Recommended Practice.  A  brief discussion on how this accuracy  is  attained in




this Recommended Practice is presented in Appendix  3.








3.1 Establishing Traceability;  Overview of  Procedures








The   analytical  procedures   described   in  this   report   for  establishing




traceability between an I/M  calibration gas and  NBS  standards  are  designed so




that traceability can  be  demonstrated  for practically any concentration  of I/M




calibration gas a  state  or  I/M customer  may  desire to  order.   The  general




procedures are  described  in the body of  this  report.  A more  concise  listing




of the procedures is provided in Appendix 2.

-------
                                      25

Generally,  the  procedures  for the  analysis of  propane,  carbon  monoxide,  or

carbon dioxide in the  I/M  calibration  gas are dependent on the characteristics

of the  analysis  instrumentation used  to determine the  concentrations  of each

gas*   A  linearity  check  is  first  conducted   to  determine  the  extent,  of

linearity betveen input gas  concentration and response in the instrument.*  If

a  four-blend  CO.,   CO,   and  propane   in . N  cylinder  is  to  be  named,  an

interference  check  is  first  conducted betveen  the  CO.  and CO.   If  only  a

three-blend  CO,  propane   and  N.  cylinder  is  to  be  made,  the  interference

check  is  omitted.   If instrument  response  during  the  linearity  check  to  a

given component in the subject gas  is  linear,  a  first  or second order equation

is developed for the instrument  from analysis  of six standard gases which have

the same  kind of  component, but  at different concentrations than  the  subject

gas.  This equation is used  in naming  the subject component  in  each of the I/M

calibration gas cylinders*  If the  instrument has a non-linear response  to  a

given  component  in  the  subject  gas,  a  third   or  fourth  order  equation  is

developed for the instrument from analysis of eight standard gases  which have

the  same  kind of  component,  but  again  at  different  concentrations than the

subject gas.  The  equation  is  then used  in determining  the concentration  of

the subject component  in each  of the I/M  calibration gas cylinders.
* Although there are  exceptions  to  the rule,  the instrumentation most commonly
used  to analyze  carbon monoxide  and  carbon  dioxide  (usually  Non-Dispersive
Infrared-NDIR-analyzer) is  non-linear, while instrumentation commonly  used to
analyze propane (usually a Flame lonization Detector-FID) is linear.

-------
                                      26

Before  presenting  details  of  the  analysis  procedures,   discussions  of  the

analysis of pure  components used  in the blending process  and  the selection of

cylinders used to hold I/M calibration gas are presented.



3.2 Analysis of Pure Components



Scientific  gas  manufacturers  must  follow certain  procedures  in  ordering  and

analyzing  pure  components  used  to  make  I/M  calibration gas.   First,  the

nitrogen used  as  diluent in the calibration  gas must be  99.99%  pure nitrogen

to comply with  this Recommended Practice.   This grade of  nitrogen  is  readily

available at a  reasonable cost and  is  frequently used  in blending  scientific

gases.  An  analysis does not have  to be  performed  on  the  nitrogen.   Propane

used  must  be   "instrument  grade"  propane  which  is certified   to  be  99.5Z

propane,  and   the   balance  will   be   other   hydrocarbons   (e.g.,   methane,

iso-butane, butane  and ethane).  An  analysis must be  performed  of  this bulk

propane  to  determine  in fact  that  the  propane gas  used is  at least  99.52

propane.  This analysis  can be  easily performed with  a  Gas Chromatograph (GC)

which  is  appropriately  fitted  to  read propane.   Lastly,   the  carbon  monoxide

used must have  a methane count of  less  than  100  ppm, and the  carbon  dioxide

used must have a total hydrocarbon (THC) count of less than 100 ppm.



Although  these  specifications for  the  pure  components  might   seem  moderately

stringent when one  considers  that the dilution of  the propane,  carbon monoxide

and  carbon  dioxide  with  nitrogen  will result  in  these  trace  components
                                      \
(particularly methane) having very  little effect on  the resultant accuracy of

the  I/M span  gas,  they  nonetheless contribute to  error in  naming  the  I/M

-------
                                      27




calibration  gas  and  should  be  limited.   Also,  the small  extra cost  of the




components ordered and analyzed  to  these  specifications will have virtually no




effect on- the overall cost of making  the  I/M calibration gas.   The bulk of the




cost in making an I/M calibration gas is  incurred in filling and analyzing the




final blend.








3.3 Cylinders








Any  size  disposable  or reusable  cylinder may be  used  as a. container  for I/M




calibration  gas  named according to  this Recommended  Practice,  however,  the




cylinders must  be constructed  of  either steel  or aluminum.   Also,  reusable



cylinders must  be  fitted with  CGA-350  valves  (CGA stands  for  the  Compressed




Gas  Association).   Disposable  cylinders  must  be  fitted  with a  CGA .1/4-inch




flare-fitting valve.   The use of these valves only will not  be a burden to gas




manufacturers,  and  they  will  allow  I/M  calibration  gas  users  to  switch




suppliers (gas manufacturers) without having to purchase new regulators.








I/M  calibration  gas  users should be  aware  that the analysis  of each  cylinder




of I/M calibration gas contributes  a  significant  portion to  the ultimate price




of  the calibration .gas.   It  probably costs  no  more   for  a manufacturer  to




perform an analysis o.f a  large cylinder than to perform an analysis  on  a small




cylinder  (except in  the case  of  calibration gas sold in disposable containers,




where  special abbreviated analysis procedures are allowed).   Therefore,  the




cost per  cubic foot of calibration  gas  sold  in  a  large  cylinder will likely be




significantly less than the  cost per  cubic foot of  calibration  gas  in. a small




cylinder.

-------
                                       28

3..4 Instrument Preparation and  Calibration


3.4.1 Definition of Linearity


The NBS standard reference materials  (either SHM's or CRH's) listed  in  Table  4-

or  a gas  divider  which  can  be  used  to  dilute  the  highest  concentration

standard  to  at least  three lover  concentrations which  match  within  ,+SZ  the

standard concentrations,  and  a instrument grade 99.99Z pure nitrogen cylinder,

must  be  used to  determine instrument linearity.   For  the purposes of  these

analytical procedures,  a linear analytical  instrument  is defined as one  which

yields three  intermediate points in  the  range of l.OZ CO  to  8.0Z CO,   250  ppm

propane  to  2500  ppm  propane  and/or  l.OZ  CO.  to  7.0Z  or  14. OZ  CO.  which

deviate by ±2Z  of point* or less  from a straight  line drawn  from the  point

determined by the zero  gas  to  the  highest calibration point.  To  be  considered

linear,   the  difference  between   the   concentrations   indicated   by   the

intermediate  points  and the   straight  line must not   exceed  ±2Z   of  the

concentration  values  of  the   intermediate  points.   The  range  of allowable

deviations from the straight line are  illustrated .in Table  4.


The reader will note  that  linearity  ia  defined  only  between  endpoints  which

coincide with available NBS standards which  are  in the  range  of interest  for

I/M calibration gases  (see  Table 4).   This  range  in which  linearity  is  defined

is wide enough  to accomodate practically any  concentration of  I/M calibration

gas ordered by a state  or I/M area.
* ±2%  of  point  is j^2Z of .reading.  For example, if  a  reading  of a certain gas
is  6.0Z CO,  +2Z  of 6.0Z  CO  is   +.122  CO.

-------
                                      29

                                    Table 4-

                        NBS Standards Used to Determine
                             Instrument Linearity*
                                                Allowable
                                                Deviations
                                                For  Linear
                   Propane in N?                Instruments

                     250 ppm                      ±5 ppm
                     500 ppm                      ^10 ppm
                     1000 ppm                     ^20 ppm
                     2500 ppm                     ^50 ppm
                   CO in N?

                      1.02                        + .022  CO
                      2.02                        +.042  CO
                      4.02                        ±.082  CO
                      8.02                        +.162  CO
                   C02 in
                     N-,

                    1.02                         * .022 C02
                    3.02                         +. .062 C02
                    4.02                         +• .082 C02
                    7.02                   •     T .142 C02
                   14.02**                       *_ .282 C02
*  An instrument-grade  99.992  pure nitrogen  cylinder must  also  be  used  to
establish the instrument zero.
**  This  C02  standard  must  be  used  in  the   linearity  check   if   the.  C02
concentration  to,,  be  named  is  over  7.02.   The  1.02  C02   standard  can  be
dropped from the check in chis situation.

-------
                                      30




Linearity  need only  be  determined  once  on a  particular  instrument,  unless




service is performed on the  instrument,  in which case the linearity check must




be conducted after service is completed.
3.4.2 C0» Interference Check
The following check is to be made only if CO  is  to be named in a cylinder that




also contains CO..  Otherwise, this step shall be omitted.








Carbon  dioxide  gas  (with  nitrogen  as  a  diluent) at  the  same  concentration




which is to be used  in the  I/M calibration cylinder must  be analyzed with the



CO analysis instrumentation.   The response of the  CO analyzer to- CO.  must be




less  than  .5Z of the  concentration of  CO which  is to  be used  in  the  I/M




calibration cylinder.








3.4.3 Calibration Curve for Linear Instruments








Prior  to the analysis of  each  batch of I/M calibration gas,  a  calibration




curve must  be generated on  the  range that is  to be used  to analyze  the  I/M




calibration gas.  No more than one  hour  must  elapse between  the  time at which




the calibration curve  is generated  and  the  start of  the analysis of  the batch




of  I/M calibration  gas begins.    If analysis of a  portion of  the  batch  is




postponed for two hours or  more,  a completely  new  calibration curve  must  be




generated.

-------
                                      31




A minimum  of six  cylinders must  be used  to  generate the  calibration  curve.




One  cylinder  must  be  instrument  grade  99.99%  pure  nitrogen,   tvo  other




cylinders  must   be undiluted  NBS  standards   which  are  above  and  below   the




concentration of  the  I/M calibration  gas to  be named,  and the  other three




cylinders can be primary standards or gases obtained  by using a gas  divider  to




dilute either  NBS standard.   Instrument responses  for the  five  points other




than the zero point should  be  approximately equally  spaced on the range  of  the




instrument which is to  be  used to name the I/M calibration gas.  For example,




if a  3.0%  CO I/M  calibration gas  is  to  named, instrument  resonses could  be




obtained at  1.0%  CO,  2.5%  CO,  4.0% CO,  6.5%  CO, and  8.0% CO.   The 1.0%  and




8.0% CO could be NBS standards, with the others being either primary standards




or diluted NBS standards.









A  first  or  second  order  equation  must  be  calculated  from  the   instrument




response for the six gases.  If the  difference between any point and the first




or second  order curve  is  less than ^1.0%'  for gases  intended to  be traceable




_+2% to NBS (_^. 5% if gases  traceable  _£!% to NBS  are to  be named  or  if CO is  to




be named  in a  cylinder containing  CO.),  the   first  or second  order equation




thereby derived may be  used to name I/M calibration  gases  in this Recommended




Practice.  Inability  to satisfy  this  criteria  with  a  first or  second  order




equation   is   an  indicator  of   improperly    named   primary   standards,   a




malfunctioning  gas divider,  and/or malfunctioning   analysis  instrumentation.




Rectifying these problems will likely result in  the criterion being satisfied.

-------
                                      32




It is suggested that  if  a  first-order equation yields a calibration line which




meets the above criteria,  but  calibration points do  not  appear to be randomly




distributed above or below the  line  (i.e.,  there is a cluster of nearby points




above the line and  another cluster below),  then a second-order equation should




be generated from  the calibration points.  This will increase  the accuracy of




the I/M calibration gas  naming  process.   The second  order  curve  must meet the




^l.OZ of point 0^.5Z if gases traceable ±IZ to  NBS  are to be named or if CO is




to be  named in  a   cylinder  containing  CO )  criterion stated in  the previous




paragraph.








3.4.4 Calibration Curve for Non-Linear Instruments








Prior to  the  analysis of each  batch  of  I/M calibration  gas  with a non-linear




instrument, a calibration curve must  be  constructed on the  range  that is to be




used in analyzing  the  I/M  calibration gas.   No more  than one hour must elapse




between the time at which  the  calibration curve is  generated and the start of




the analysis  of the  batch of  I/M  calibration gas  begins.   If  analysis  of  a




portion of  the batch  is postponed  for  two  hours  or  more,  a completely  new




calibration curve must be generated.








A minimum of  eight  cylinders must be used  to construct the calibration curve.




One  cylinder  must  be   instrument-grade   99.99%   pure  nitrogen,  two  other




cylinders  must  be undiluted  NBS  standards  which  are  above  and  below  the




concentration  of the  I/M calibration  gas. to  be  named, and  the other  five




cylinders can be primary standards or gases obtained by using a gas divider to




dilute  either  NBS   standard.  Instrument  responses  for the  seven points other

-------
                                       33

 than  the  zero  point should be approximately  equally  spaced on the range  which

 is to be used  to name  the  I/M  calibration  gas.



 A  third  or  fourth-order  polynomial  equation  must  be  calculated  from  the

 instrument responses to  the eight  gases obtained during  the calibration  step.



 No inflection  points are  allowed  in  the  equation of  the  curve generated  from

 analysis of  the eight  gases.*   If  an inflection point occurs in a  fourth-order

 equation, a  third-order equation  should be  tried.   If the differences  between

 any point and  the  curve is less than ^l.OZ  for  gases  intended to be  traceable

 _+2% to  NBS  (^.5Z  if gases  traceable ^11 to NBS are to be  named or if CO  is to

 be named in a cylinder containing  CO.),  the  equation thereby  developed  may

 be used  to name  I/M calibration gases  in  this Recommended  Practice.   Inability

 to  satisfy  these  criteria  with  a  third   or   fourth-order  equation  is  an
                         \
.indicator of improperly named  primary standards,  a malfunctioning gas divider,

 and/or malfunctioning  analysis instrumentation.
 * Inflection points can be  determined  by taking the  second  derivative of  the
 resultant  calibration  curve  equation,  setting it equal to zero, and evaluating
 over .the range  of  0  to 100%  full  scale  on the  given  range.

-------
                                      34




It  is  suggested  that if  a third-order  equation yields  a  calibration curve




which meets  the  above  criteria  for non-linear  instruments,  but calibration




points do  not  appear to  be  randomly distributed  above or below  the curve




(i.e.,  there is  a cluster of  nearby  points  above the curve  and another below




the  curve),  then  a  fourth-order  equation  should  be  generated  from  the




calibration points.   This will  increase the  accuracy of the  I/M calibration



gas  naming  process.   However,   no  inflection  points   are  allowed  in  this




fourth-order curve.   The  fourth-order  curve must meet  the ±IZ of point (^.51



if gases traceable ±1!  to NBS are  to be named  or if CO  is  to be named  in a




cylinder containing CO.) criterion stated in the previous paragraph.








3.5 Analysis of I/M Calibration Gas Cylinders








3.5.1 Re-Usable Cylinders








After valid calibration curves are  obtained  for all  instruments that  are to be




used to analyze  the  I/M calibration gas, the  analysis of I/M  calibration  gas




can  begin.  Each component  (with  the  exception  of  the  diluent)  of  every




re-usable  I/M  calibration gas  cylinder must  be discreetly  analyzed.   I/M




calibration  gas  cylinders can  be  analyzed  -in  any sequence,  however,   the




highest concentration NBS  standard  used1 to generate  the  calibration  curve  and




the  zero   gas  cylinder  must   be  repeated  after  -every  ten. analyses  of  I/M




calibration gas  cylinders  to  ensure  that  substantial instrument  zero  or  span




drift has  not occurred  which  could  detrimentally  affect the  I/M calibration




gas  naming process.   No analyzer adjustments  aye permitted  during the  naming




process.    Zero and span  drifts are  calculated as a percentage  of  full  scale.

-------
                                      35




They are added together  to  arrive  at a total drift error.   If the total drift




error changes by more than^l.OZ (_+.5Z if gases  traceable'^1Z to NBS are to be




named or if  CO  is to be  named  in  a cylinder  containing CO   to _+1.5Z  to NBS)




throughout the analysis  process, a  new  calibration curve  must  be constructed




for the analyzer  in  accordance  with the procedures described in section 3.4.2




or 3.4.3,  and  the analyses  of  the previous ten cylinders  of  I/M calibration




gas must be repeated.








3.5.2 Disposable Cylinders








It is likely that much of the I/M calibration gas  sold  to I/M users will be in




the form of  disposable cylinders.   Many  analyzer manufacturers  are  designing




the  stands that  hold  their  analyzers  to  include a holder  for  a disposable



cylinder that  can  be used  to  periodically  calibration-check  the  analyzer.




These disposable  cylinders  hold a small  amount  of calibration  gas,  usually 6




to  10  cubic  feet.    The   disposable   cylinders  are  likely   to  be  used




predominantly  in licensed   facilities  of  decentralized programs,  where  (l)




there may  be no  room for a  large  cylinder  of  I/M calibration gas, or  (2)  the




facility owner  does not  want to  incur  the higher  initial  cost  of a  larger



cylinder.   Fewer  vehicles will  be  inspected on  the  basis  of each disposable




cylinder.








As  stated  previously,   for re-usable   cylinders  the  Recommended   Practice




requires that an  analysis be performed according  to  the prescribed techniques




on  each component  (with the exception  of the  diluent)  in every  cylinder.




However, the result  of  insisting on this requirement for disposable  cylinders

-------
                                      36




also might  be  to  make them  prohibitively expensive  for small  facilities in




decentralized programs.    .








Disposable cylinders  are  usually  filled  in one of  two ways: (1)  they can be




filled  one-by-one  from a  re-usable cylinder  or bulk  homogeneous  mixture, or




(2) separate cylinders  of  pure components can  be  piped together and connected



to  a  common  manifold  to  which  the  empty   disposable  cylinders  are   also




connected,  thereby  filling   the  disposable  cylinders  simultaneously   from  a



single stream of "dynamically" blended  gases.   In this latter filling process,




the blend is  usually continuously analyzed while  the  disposable cylinders are




being filled.   In  the first  filling  process  mentioned, all  components of the




mixture  have  usually  been   analyzed  already.   In   both   filling  processes




mentioned, analysis is  usually performed  on the gas blend somewhere along its



path to the disposable  cylinder, which  makes  the probability of  the mixture in




the disposable cylinder being  accurate  much greater  than if  an analysis of the




blend  entering  a  disposable  cylinder  had  not  been   performed.   However,




blunders  (cylinders   with  contents  significantly  different  than  the  labeled




concentrations)  can   still  occur  with  the disposable  cylinders,  particularly




where a  disposable cylinder is not properly evacuated prior to  filling,  or  a




disposable cylinder is  not properly filled because of  some  malfunction in the




ganging hardware.








For these reasons, this Recommended  Practice requires  only that  an analysis be




performed on  (1) one  component of each filled disposable  cylinder,  and on (2)




the other  components (with  the exception of  the  diluent)  in 10% and  no. less

-------
                                      37

than ten*  total cylinders of  the filled  disposable  cylinders that  have been

blended with a  continuous  analysis system or filled  from a previously blended

and  analyzed bulk mixture.    The mean  concentration of  the  first  component

which is analyzed  in  all  disposable  cylinders in the batch is calculated.  If

the concentration  of  the  first component is any cylinder  in  the  batch is more

than ±2%  different from  the mean concentration, that  cylinder must  have all

components   discreetly   analyzed.   Any   cylinder  whose   first   component

concentration is within ±2%  of the mean  concentration  of  the  batch  may apply

the  criterion  of the  following  paragraph  in  the   determination  of  the

concentration of other  components  in that cylinder.



The  mean  concentration  of  the   other  component  in the 10Z  sample(s)  are

calculated.  The  standard deviation of  the concentration values  of  the other

components" in  the  10Z samples must be  less than or  equal  to  i»5Z  of each

sample's mean.**   If  they are-, all cylinders  in the batch  shall be  labeled

with the mean  concentrations of the other  components  in  the  10Z  samples.   If

they  are  not,  an analysis  of all  cylinders  for "the   other  components  is

required.
* If less than 10 disposable  cylinders  are  to  be  named,  all components (except
the diluent) must be analyzed in each cylinder.
**  This  will assure  that 95Z  of  the  sample  will be within  ±1% of  the  mean
concentration (^2 standard  deviations), and ?9Z  of  the  sample will  be  within
^2% of the mean  concentration (±4  standard  deviations).   For an explanation of
this statistical criteria see Appendix 4.

-------
                                      38




If a continuous  analysis  system for all  components  (propane, CO  and  CO ) has




not been  used in the  case  where  disposable cylinders  are filled  from pure




components  flowing  through a  common manifold,  or where  an analysis  has not




been performed  on  each  component  in the  bulk  mixture  in  the case  where




disposable  cylinders  are  filled  from the  bulk mixture,  an analysis  must  be




done on each component of every disposable cylinder.








3.6 Calculating I/M Calibration Gas Concentrations








3.6.1 Equations Used








Linear Instruments - The concentration of each component of an I/M calibration




gas (propane and  CO)  that  has  been analyzed with  a  linear  instrument  shall  be




determined by  the first  or second order  equation  generated by  the calibration




curve as discussed in Section 3.4.2.








Non-Linear  Instruments  -  The  concentration of  components  in  I/M calibration




gas analyzed with a non-linear instrument shall be calculated by  the  third  or




fourth-order calibration curve equation discussed  in section 3.4.3.








3.6.2 Concentration Determination








Re-Usable Cylinders - Propane, carbon monoxide,  and  CO. if  used  are  analyzed




in  each  re-usable  cylinder   of  I/M  calibration  gas   blended.   The  labeled




concentration  of  each component in each  cylinder must be  calculated  from the




meter  responses  for  each  component using  the  equations specified in Section




3.6.1.

-------
                                      39




Disposable Cylinders  - For  disposable  cylinders  in  which each  component has




been analyzed in all  cylinders  of a batch (see  Section  3.5.2),  the same rules




apply for determining- the concentration  label  of each component  as  apply for




re-usable cylinders  (see  previous  paragraph).   For  disposable  cylinders  in




which one component  has  been  analyzed  in every  cylinder,   while  the  other




components have bean  analyzed in 10Z (and not  less  than 10)  of the  cylinders




of  the   batch,   the  concentration  label of   the  first  component  shall  be




calculated  separately  from   the  meter   response of  each  cylinder  using  the




equations specified in Section 3.£.1.   The  concentration label  for  the  other




components on all cylinders  shall  be the mean concentration values of  the LOZ




samples, provided all other conditions in Section 3.5.2 have been met.








3.7 Cylinder Labeling and Documentation








All I/M calibration gas  cylinders  named according to  these procedures  must  be




labeled  with a tag which contains at a minimum the following information:








    (i)    Cylinder  number, except  in the use  of disposables,  where  the  batch



          number is required




    (ii)  Concentration of propane  (in  ppm),  CO  (in  molZ),. and  CO.  (in  molZ)



          in  cylinder  gas   (determined  from  Section  3.6.2),   and   accuracy




          specification (i.e., ^22, +1.5Z or ^l.OZ)




    (iii) Balance gas




    (iv)  Analysis  date



    (v)    Cylinder  numbers of  NBS  standards  used in determining  instrument




          calibration curves

-------
                                      40



    (vi)  Vendor name
                ' /


    (vii) The statement  that "This gas has  been named  in  accordance with the


          EPA Recommended Practice for Naming I/M Calibration Gas".






The gas manufacturer must retain calibration curve  data on each batch analysis


of I/M calibration gas for a minimum period of two years.






3.8 Audits of I/M Calibration Gases






A formal  audit  of  I/M calibration gas which has been named  according  to this


practice  is  not  necessary  prior  to  the  sale  and  delivery  of  the  I/M


calibration  gases.   However, it should  be  remembered  that  state  auditors  in


decentralized  I/M. programs  will likely  be  checking  on a  monthly  basis  the


accuracies  of calibration  gases used with  inspection  analyzers.   Facilities


with improperly  labeled  calibration gases will  likely  be  required by auditors


to  suspend  the  conducting  of inspections  until  the  problem  with  their


calibration  gas  is  resolved.    The  labeling requirements  of  the  Recommended


Practice  should  assist all  concerned parties in determining  why  a cylinder  or


group of  cylinders was improperly labeled.






The  EPA  may,  from  time-to-time,  conduct  audits  of  I/M  calibration  gas  by


acquiring  cylinders  named  according  to  the  Recommended Practice on  the open


market  and  analyzing these cylinders.  The  results  of  such  audits  are  likely


to be published  by the EPA for the benefit of I/M programs.

-------
    41
APPENDICES

-------
                   41
           Appendix  1

 Emission Performance  Warranty
 Regulations Pertaining to I/M.
          Calibration Gases
§ 8&2217  Calibrations, adjustments.
  (a) Equipment shall be calibrated in
accordance with the manufacturers'
instructions.
 • (b) WHhin one hour prior to a test, the
analyzers shall be zeroed and spanned.
Ambient air is acceptable as a zero gas:
an electrical span check is acceptable.
Zero and span checks shall be made on
the lowest range capable of reading the
short test standard.
  (c) Within eight hours prior  to a
loaded test, the dynamometer shall be
checked for proper power absorber
settings.
  (d)(l) The analyzers shall have been
spanned and adjusted, if necessary.
using gas traceable to NBS standards ±
236 within one week of the test. These
span gases shall have concentrations
either
  (i) Between the standards specified in
this subpart and the jurisdictions
inspection standards for 1981  model
year light duty vehicles, or
  pi) Be within -50% to +100% of the
standards in this subpart.
  (2) For analyzers with'a separate
calibration or span port, CO readings
using calibration gas through  the probe
and through the calibration port shall be
made: discrepancies of over 33S shall
require repair of leaks. No analyzer
adjustments shall be permitted during
this check.
                           40 CPS. Part  85
                           Page  34308
                           May 22,  1980

-------
                                      42




                                  Appendix 2








                       Listing of Analysis Requirements




                           for Recommended Practice
Step 1       :  Analyze pure propane used to blend I/M calibration gas for




Analysis        other hydrocarbons such as methane, iso-butane, butane and




of Pure         ethane.  Analyze pure carbon monoxide for methane.  Analyze




Components      the  pure   CO.   (if  used)  for   total   hydrocarbons.    These




                analyses are not necessary  prior  to  the  blending of each batch




                of I/M calibration gas, but should be performed sometime prior




                to the  blending of  the  first  I/M calibration gas  batch  made



                from newly received pure components.








     Acceptance Criteria: The total "other" hydrocarbon count of pure propane




                (i.e., excluding propane)  must be less than  .5%.   The  methane



                count of pure carbon  monoxide  must be less  then  100 ppm.   The




                total hydrocarbon count of  the CO- must be less than 100 ppm.
Step 2       :




Determination




of Insrument




Linearity
Determine whether instrument(s) to be used in analysis of




I/M calibration gas is (are) linear or non-linear.   The




following NBS standards (either CRMs or SRMs) must  be used in




determining  linearity,  or  a gas  divider  may be used  with  the




highest  concentration  standards  (i.e.,  2500  ppm  propane  in




N2,  8.0%  CO in  NZ  or  14.0%  C02  in  Nj)  listed  below  to

-------
                      43




obtain three  points  below the  highest  concentration standard.




If a gas  divider  is  used, the  lower  points  should be targeted




to within ±5Z of  the concentration of  the  standards listed in




the table.   An instrument grade  99.99Z pure nitrogen  must be




used  to  establish   the  instrument   zero.    The  14.01  CO.




standard  must  be used   in  the  linearity  check  if  the  CO.




concentration  in  the I/M calibration gas  to be  named  is over




7.0Z  CO..   The  l.OZ  CO.  standard   can  be  dropped from  the




check in  this  situation.

-------
                                  44
                   NBS Standards  Used  to Determine
                         Instrument Linearity
                                          Allowable
                                          Deviations
                                          For Linear
             Propane in N?                Instruments

              250 pptn                      ^ 5 ppm
              500 ppm                      ^10 ppm
              1000 ppm                     ^20 ppm
              2500 ppm                     ^50 ppm
             CO in N-?

              l.OZ                         + .02Z CO
              2.02                         ,+.04Z CO
              4.0Z.                         "+.08Z CO
              8.0Z                         ™.16Z CO
              l.OZ                        ^.02Z C02
              3.0Z                        ^.06Z C02
              4.0Z                        +.08Z C02
              7.0Z                        ^.14Z C02
             14.OZ                        + .28Z C02
•Acceptance  Criteria:   The criteria for determining instrument linearity

             is as follows.  A  linear analytical instrument  is  defined as

             one which  yields  three  intermediate  points  in  the  range of

             l.OZ CO  to  8.0Z  CO,   250  ppm  propane  to  2500 ppm  propane,

             and/or   l.OZ  CO.  to  7.0Z  CO.  or  14. OZ  C02J   (depending on

             the CO.  concentration to  be  named) which  deviate  by ±2% of

             point or   less  from  a  straight  line  drawn  from  the  point

-------
                                      45




                 determined by  zero  gas to the highest  calibration point.  To




                 be    considered    linear,    the   difference    between   Che




                 concentrations  indicated  by Che  intermediate points  and the




                 straight line must  not exceed' _±2Z of the concentration values




                 of   the  intermediate   points*    The   range   of  allowable




                 deviations from  the  straight  line is illustrated in Che Cable




                 above.  Instruments  not meeting  this criteria  are classified



                 as non-linear instruments  for the purpose  of Chis Recommended




                 Practice.








Seep 3;       :  The following check is Co be made only if CO is to be named




CO-              in an I/M calibration cylinder Chat also contains CO.,




Interference     otherwise Chis step shall be omitted.




Check








                 CO.   gas   (with  nitrogen  as   a  diluent)   at  Che   same



                 concentration  which is  to be  used  in  Che I/M  calibration




                 cylinder   must   be    analyzed   with    Che    CO   analysis




                 instrumentation.








     Acceptance CriCeria;  The response of Che CO analyzer Co CO^ must  be




                 less  Chan .52 of Che  concencration of CO which  is  Co  be  uaed




                 in Che  I/M calibration cylinder.

-------
                                      46
Step 4;




Generation




of




Calibration




Curve for




Linear




Instruments
Prior to the analysis of each batch of I/M calibration gas, a




calibration curve must be generated on the range that is to




be used to analyze the I/M calibration gas.  No more than one




hour must elapse between the time at which the calibration




curve is generated and the start of the analysis of the batch




of calibration gas begins.  If analysis of a portion of the




batch is  postponed for  two  hours  or  more,  a  completely new




calibration curve must be generated.
                 A  minimum of  six  cylinders must  be  used  to generate  the




                 calibration  curve.    One  cylinder  must  be  instrument  grade




                 99.99% pure  nitrogen,  two  other  cylinders must be undiluted




                 NBS standards which  are  above and below  the  concentration of




                 the  I/M  calibration  gas to be  named,   and  the other  three




                 cylinders can be primary  standards  or  gases  obtained by  using




                 a  gas divider  to  dilute  either NBS  standard.   Instrument




                 responses  for  the  five  points  other  than  the   zero  point




                 should be  approximately  equally  spaced  on  the range of  the




                 instrument which  is  to  be  used  to  name  the  I/M  calibration




                 gas.  For  example,  if  a 3.0%  CO I/M calibration  gas  is  to




                 named, instrument resonses  could  be obtained  at 1.0% CO, 2.5%




                 CO,  4.0% CO, 6.5%  CO,  and 8.0% CO,   The  1.0%  and  8.0%  CO




                 could be  NBS standards,  with the others  being  either primary




                 standards or diluted NBS standards.

-------
                                 47




            A first  or  second order equation must  be calculated from  the




            instrument response for the six gases.








Acceptance Criteria; If the difference between any point and the first  or




            second order  curve is less  than ±1.0Z  for  gases  intended to




            be traceable ^22  to NBS  (+.5Z  if gases  traceable  +IZ  to  NBS




            are  to  be  named  or  if  CO  is  to   be  named  in.  a  cylinder-




            containing CO.),  the first  or  second order equation thereby




            derived  may  be used to  name I/M calibration gases  in this




            Recommended  Practice*   Inability  to  satisfy  this criterion




            with  a  first  or  second  order  equation  is an indicator  of




            improperly  named  primary  standards,   a  malfunctioning   gas




            divider,   and/or   malfunctioning  analysis   instrumentation.




            Rectifying these  problems  will  likely result  in  the criteria




            being-  satisfied.








            It  is  suggested   that  if  a  first-order equation yields a




            calibration   line  which  meets   the   above   criteria,    but




            calibration points do  not  appear to be randomly  distributed



            above  or below the line  (i.e.,  there is  a  cluster of nearby




            points  above   the line  and  another  cluster  below),  then a




            second-order   equation   should   be    generated   from    the




            calibration points.   This will  increase the accuracy of  the




            I/M  calibration gas naming  process.   The second  order  curve




            must meet the  _^1.02 of point C^.5Z if  gases traceable  ^12 to




            NBS  are  to be  named or  if  CO  is to be named in  a  cylinder




            containing CO.) criterion stated in the  previous paragraph.

-------
Step 5;




Generation




of




Calibration




Curve for




Non-Linear




Instruments
Prior to the analysis of each batch of I/M'calibration gas




with a non-Linear instrument, a calibration curve must be




constructed on the range that is to be used in analyzing the




I/M span gas.  No more than one hour must elapse between the




time at which the calibration curve is generated and the




start of the analysis of the batch of I/M calibration gas




begins.  If  analysis of a  portion of the batch  is postponed




for  two  hours or more,  a  completely  new calibration  curve




must be generated.
                 A minimum of eight  cylinders  must  be  used to  construct  the




                 calibration  curve.    One  cylinder  must  be  instrument-grade




                 99.99Z pure  nitrogen,  two other  cylinders must  be  undiluted




                 HBS standards which  are  above  and below  the  concentration of




                 the  I/M   calibration gas  to  be  named,  and  the other  five




                 cylinders can be primary  standards or gases obtained by using




                 a  gas divider  to  dilute either NBS  standard.   Instrument




                 responses  for the  seven points  other  than  the zero  point




                 should be  approximately  equally spaced on  the range  which is




                 to be used to name the I/M calibration.gas.
                 A   third   or   fourth-order  polynomial   equation  must   be




                 calculated from  the  instrument  responses  to the  eight  gases




                 obtained during the calibration step.

-------
                                 49




Acceptance Criteria; No inflection points are allowed in the equation  of




            the curve generated  from analysis  of the eight gases.*  If an




            inflection  point   occurs  in  a   fourth-order  equation,  a




            third-order  equation  should  be  tried.   If  the  differences




            between any point  and  the curve is  less  than  ±1.0Z for gases




            intended to be  traceable ±2Z to NBS (+.5Z  if gases traceable




            ±IZ  to  NBS are  to be  named  or  if CO  is  to  be  named  in a




            cylinder  containing  CO.),  the  equation  thereby  developed




            may be used to  name  I/M calibration gases in this Recommended




            Practice.  Inability to satisfy these criterion with  a third



            or fourth-order  equation is an indicator of improperly named




            primary  standards,  a   malfunctioning  gas  divider,   and/or




            malfunctioning analysis  instrumentation.








            It  is  suggested  that   if  a  third-order  equation  yields  a




            calibration   curve  which  meets   the  above   criteria  for




            non-linear instruments,  but calibration points do  not appear




            to be randomly  distributed  above  or  below the curve (i.e.,




            there  is  a   cluster  of  nearby points  above  the   curve  and




            another below the  curve),  then  a  fourth-order  equation should




            be generated  from  the  calibration  points.   This will increase




            the  accuracy  of  the   I/M calibration  gas  naming  process.




            However,   no   inflection  points   are    allowed   in    this




            fourth-order  curve.    The  fourth-order curve  must meet  the




            +IZ  of  point  (*.5Z  if  gases  traceable j*lZ  to  NBS   are  to be



            named or  if  CO  is to be named  in a cylinder  containing  CO )




            criterion stated in  the  previous paragraph.

-------
                                      50
Step 6A



Analysis of



Re-Usable



I/M



Calibration



Gas Cylinders
Steps 5A and SB differ in that Procedure A is  for analysis of



re-usable cylinders, and Procedure B is for analysis of



disposable cylinders.







After valid calibration curves are obtained for all



instruments that  are to be  used to analyze  calibration gas,



the  analysis   of  I/M  calibration  gas  can  begin.   Each



component  (with  the  exception  of  the  diluent)  of  every



re-usable  I/M  calibration  gas  cylinder  must be  discreetly



analyzed.  I/M  calibration  gas  cylinders  can be  analyzed  in



any sequence, however, the highest  concentration NBS  standard



used  to  generate   the  calibration  curve  and  a  zero  gas



cylinder must  be  repeated  after  every ten  analyses of  I/M



calibration   gas   cylinders   to   ensure   that   substantial



instrument zero or  span  drift has  not occurred which • could



detrimentally affect  the  I/M calibration  gas  naming  process.



No  analyzer-  adjustments  are  permitted  during  the  naming



process.  Zero  and span drifts  are  calculated  as a percentage



of full  scale.  They are  added together to arrive  at  a  total



drift error.
     Acceptance Criteria; If the total drift error changes by more than ^1.0%




                 (_*. 5Z if gases  traceable ,+lZ  to NBS are to be  named  or if CO




                 is  to  be  named  in  a 'cylinder containing  CO.  to ^1.52  to




                 NBS)  throughout the  analysis  process,  a new calibration curve




                 must  be constructed  for the analyzer  and  the  analyses  of the


                                \i

                 previous  ten   cylinders   of   I/M   calibration  gas  must  be
                                 \



                 repeated.

-------
                                      51
Step 6B




Analysis of




Disposable




I fli




Calibration




Gas Cylinders
After valid calibration curves are obtained for all




instruments that are to be used to analyze the I/M




calibration gas, the analysis of I/M calibration gas can




begin.  For disposable I/M calibration gas cylinders, the




requirement is that an analysis be performed on (l) one




component of  each  filled disposable cylinder, and  on  (2) the




other components,  (with  the  exception  of the  diluent)  on only




10Z  and  no  less  than  ten  cylinders  total of   the  filled




disposable cylinders that have been  blended with  a continuous




analysis  system  or   filled   from  a  previously  blended  and




analyzed bulk mixture*
                 The  mean   concentration  of  the  first   component   in  all




                 cylinders in the batch is calculated.








                 The  highest concentration NBS  standard  used to  generate the



                 calibration curve  and a zero gas  cylinder must  be  repeated




                 after  every ten  analyses  of  Che  component which is  being




                 analyzed  in 10OZ  of the  disposable  cylinders.   No  analyzer




                 adjustments are permitted during these analysis processes.
     Acceptance Criteria;  If the concentration of the first component in any




                 cylinder  in the  batch is  more than _£2Z different  from the




                 mean  concentration,  that  cylinder must  have  all  components




                 discreetly analyzed.

-------
                     52


If  the  total drift error  does  not change by  more than ^1.0%


(±.5Z if  gases  traceable ^12 to NBS are to  be  named  or if  CO


is  to  be named  in a.  cylinder  containing  CO.  to ^1.52   to

                                   »
NBS)  throughout the analysis  process,  the  process  is valid.


Changes which exceed the  criteria above  require that  a new


calibration  curve must  be constructed  and  the  previous ten


cylinders be re-analyzed.




For  the other components  in which  only  10Z  of  the  batch has


been  analyzed,  the  standard deviation  of  the  concentration


values  of the  other components  in the  10Z samples  must  be


less  than or equal  to  ^.52 of  each  sample's  mean.   If they


are,  all  cylinders  in  the  batch shall  be  labeled with the


mean  concentrations  of  the other component  in the respective


101 samples.  If  it  is  not, an  analysis  of  all cylinders for


the other components is required.  If  less than 10 disposable


cylinders are to  be named, all  components except  the  diluent


must  be analyzed  in each  cylinder.   Also,  if  a continuous


analysis  system for each  component -has  not  been  used in the


case  where  disposables  are  blended  from  pure  components


flowing through a common manifold, or  where an  analysis  has


not  been  performed on each  component  in the bulk mixture  in


the case  where  disposables are  filled from  the  bulk  mixture,


an  analysis  must  be   done  on  each   component of  every


disposable cylinder.

-------
                                      53
Step 7A




Determination




of




Concentration




Values for




Re-Usable




Cylinders
Steps 6A and 6B differ in their applicability for either




re-usable or disposable cylinder.








The labeled concentration of each component in each re-usable




cylinder must be calculated from the meter responses for each




component using the equations developed for the calibration




curves.
Step 7B




Determination




of




Concentration




Values for



Disposable




Cylinders
For disposable cylinders in which each component has been




analyzed in all cylinders of a batch, the same rules apply




for determining the concentration label of each component as




apply for re-usable cylinders.  For disposable cylinders in




which one component has been analyzed in every cylinder,



while the other components have been analyzed in only 102 of




the  cylinders  of the  batch,  the  concentration label  of  the




first  component  shall  be  calculated  separately  for  each




cylinder  using the  equations developed  for the  calibration




curves.   The concentration  label  for the other  components  in




all  cylinders  shall be the mean  concentration values  of  the




10Z  samples, provided  all other conditions  pertaining  to  the




10%  samples have been met.

-------
                                      54





Step 8        :  All I/M calibration gas cylinders named according to these




Cylinder         procedures must be labeled with a tag which contains at a




Labeling         minimum the following information:




and




Documentation









              (i)    Cylinder number,  except  in the use of  disposables,  where




                     the batch number is required




              (ii)   Concentration of  propane (in ppm)  CO  (in molZ),  CO   (in




                     mol Z)  in cylinder gas  (determined from  Section  3.6.2),




                     and accuracy specification (i.e., ±2Z, H.5Z or^l.OZ)




              (iii)  Balance gas




              (iv)   Analysis date




              (v)    Cylinder  numbers  of  NBS  standards used  in  determining




                     instrument calibration curves




              (vi)   Vendor name




              (vii)  The statement that  "This gas  has  been  named  in accordance




                     with EPA  Recommended  Practice for  Naming  I/M Calibration




                     Gas".








                 Fhe  gas  manufacturer must  retain calibration  curve  data  on




                 each  batch analysis  of  I/M calibration  gas  for  a  minimum




                 period of two years.

-------
                                      55

                                  Appendix 3

                              Accuracy Discussion



The purpose  of this discussion  is  to explain how the accuracy specifications

for I/M  calibration gases to NBS  are derived within  the Recommended Practice

for I/M  calibration gas  filled  in a  re-usable  cylinder.   The  discussion for

disposable cylinders would be very similar  although it  would  include a small

statistical  sampling error  for  components  which  are not  analyzed  in every

disposable cylinder.
                   ,•


At  certain places  in  the  Recommended Practice  small working  tolerances  are

allowed  which  help make  the   process   of  naming I/M  calibration  gases ' a

relatively smooth process  without  sacrificing  a  great  deal of accuracy.   These

tolerances  are  limited  to  certain   amounts  so   that   the  overall  accuracy

specifications can be maintained.



The maximum total error or uncertainty  in a  given I/M calibration cylinder can

be  estimated  by  adding  together  the maximum  tolerances  allowed  from  each

source of  error.*   Of  course, gas manufacturers  in many cases may not  use—up

the maximum  tolerances  allowed  in this Recommended  Practice.   But  if even the

worst case  stack-up of tolerances meets  the required  accuracy specifications,

all cylinders in a batch should be ^22 to NBS or better.
* A certain maximum percentage  error allowable in the naming  process  does not
always translate to the  same percentage error in  the  labeled  concentration of
a cylinder.   In most cases  the  error of the  labeled  concentration value would
be  less  than the  error allowed  in  the  naming,  process.  However,  for  the
purpose  of this  analysis,  we  have  assumed  that  a  certain  percentage  error
allowed  in the  naming process  translates to  the  same percentage  error  in the
labeled  concentration value of  a cylinder.

-------
                                      56

Table .5  presents  an accuracy  analysis  of  two  different  types  of calibration

gas mixtures theoretically named  by  the Recommended Practice.  These gases are

representative of what most  I/M programs will be  using.   The error tolerances

for each source of -error in  the Recommended Practice are added to determine an

overall accuracy  (bottom  of  the table), which  can be compared  to the desired

accuracy specifications at the  top of the table.
                                   .Table 5

                           Accuracy Analysis of Two
                      Common I/M Calibration Gas Mixtures
                  Named to Different Accuracy Specifications
                         With the Recommended Practice

                     Accuracy Spec Desired in   Accuracy Spec Desired in
                     I/M .Calibration Gas        I/M Calibration Gas
                     of Propane, & CO in N2     of Propane, CO, & C02 in
Error Sources

(1) CO2 Interference

(2) Calibration Curves

(3) Total (Zero and
    Span) Drift
+22 to
NBS
N/A*
I1*
+12
+12 to
NBS
N/A*
+.52
+.52
+1.52 to
~~ NBS
+.52
+ .52
+.52
+22 to
NBS
+.52
+..52
+ 12
    Overall Accuracy
 (Total of 1, 2, and 3)
+22
+ 12
+ 1.52
+22
* Not appliable because this gas mixture does not contain C02

-------
                                     • 57


The  reader  will  note  that  an accuracy  specification of  ±IZ  to  NBS  is not


available in  the Recommended  Practice for  an I/H calibration  gas containing


CO   and  CO.   This  is  because  the  added  error of  the  interference between


CO   and  CO  cannot  be  reasonably  compensated for  by lowering  the acceptable


tolerances of the other sources  of error  (i.e.,  calibration curve and zero and


span drift error).  To do so would make the  process  of  naming I/M calibration


gases unduly restrictive and expensive.





Many  states  have  specified that  their auditors  shall  use  calibration gases


which are traceable ±1Z to  NBS,  while  the  inspectors  shall use gases which are


traceable ±2Z to NBS.  The  reasoning was that audit gas  being used to pass and


fail the inspector's gas should be more accurate then the inspector's gas.





From  Table  5  it  is  apparent   that  auditors  are  not  assured  with  this


Recommended  Practice  of  obtaining a  four blend  CO  ,  CO,  and  propane  in  N


audit gas that  is ±1Z to NBS.   Some  gas  manufacturers,  however, may  have the


capability to name all components  of  a four—component mixture to within ^IZ of


NBS  standards.   The best  specification within  the  scope  of  the  Recommended


Practice  for  the four blend mixture  containing  CO and  CO- mixture  is  H.5Z.


Auditors  either  can use  this gas  or,  if  they desire  £LZ gases, may  (1)  carry


two  cylinders  which are  ^12  to  NBS,  one being  CO. in  N_,  the  other  being


CO  and  propane  in N,,  or  (2)  contact   a  scientific  gas  manufacturer  to
                                 ' r                              •

discuss whether or not a  four  component, gas mixture named  ^1% to NBS standards


can be supplied.

-------
                                      58




                                  Appendix  4
                      Discussion of Statistical  Sampling




                   and Naming Procedure  for I/M  Calibration




                     Gases Filled in Disposable  Cylinders








The  Recommended  Practice  uses  statistical  sampling criteria  to  determine




whether or  not an abbreviated  procedure can  be used in  analyzing  and naming




I/M calibration gases  filled  in disposable cylinders.  An example is presented



below.








A  batch  of one-hundred  disposable cylinders  filled  with  2.0% CO,  5.02 CO.,




and 600  ppm propane in  N_ is  to  be  analyzed and  named.   The  batch  has been




filled simultaneously  from a previously analyzed bulk mixture.   The following



procedures illustrate the  requirements of  the  Recommended  Practice.








All one-hundred  cylinders are  analyzed  for CO  concentration.   The  CO  labeled



values  are  determined  by the analyzer  meter  response  to each  disposable




cylinder  and  the  third   or  fourth order   equation  developed  for  CO.   A  10%




sample (with no less than  10  cylinders)  is randomly selected from the batch of




100  cylinders  for  further analysis  of CO.  and  propane.   If  the  batch had




consisted of  less than  one-hundred cylinders,  ten cylinders would  still have




to  be selected.   For  batches  larger than one-hundred cylinder,  10%  of the




batch  is  required.    The  ten  cylinders  are  analyzed for  CO.  and  propane

-------
                                      59



concentrations.   These concentrations are  calculated from  the analyzer  meter



responses  to  the cylinders  and  the equations developed  for  each  analysis



system.







The mean CO  and propane  concentration of  the  sample is  determined.   If  the



mean  concentration  values  for  C0_ and  propane in  the  subject  sample are  to



be  used for  the  entire batch,  the  sample  must display  a certain  amount  of



homogeneity.   This  is established  by setting limits  or  tolerances around  the



meac  concentration,  outside of which the standard  deviation  of  concentration



values must not vary if homogeneity is assumed.






The  tolerance limits  set  by  the Recommended  Practice are ^.52  of  the mean



concentration.  If  the standard deviation is  within these  limits,  one can  be


a-ssured  that  66Z  of  the  batch is  within i«5Z  of  the mean concentration  (±1



standard deviation),  952 of  the batch is within £LZ  of  the mean  concentration



0^2 standard  deviations) and  somewhat over 99%  of the batch is within ^22  (+4



standard deviations)  of  the mean  concentration.   If the mean concentrations  of



CO-  and  propane  are  4.982  CO   and 603  ppm,  the  tolerances   around   these



values  are  4.982 CO.  ^.0252  CO.,  and   602  ppm  .+3  ppm.    If   the   standard



deviation  of CO-  and propane  concentration values  are • within  these  limits,


                                                 ' **         * A
all cylinders  in  the  batch can be labeled with  these mean concentrations.   If



the standard  deviations  lie  outside of these  tolerances,  all  cylinders in  the



batch  must  be analyzed   for   CO   and  propane, the concentrations   of  each



cylinder  being  calculated from the  analyzer meter responses  to  each  cylinder



and the equation developed for each analysis system.

-------