EPA-AA-SDSB-80-
                              Technical Report

                               February, 1980
                         Recommended Specifications
                        for I/M Type Analysis Systems
                        Used to Pass or Fail Vehicles
                                     by

                             William B. Clemmens
                              Merrill W. Korth
                              Gordon J. Kennedv
                                   NOTICE

Technical Reports do  not  necessarily represent final EPA decisions or posi-
tions.  They are intended  to present technical analyses of issues using data
which are currently  available.   The purpose  in the  release  of such reports
is  to  facilitate  the  exchange  of  technical  information and  to  inform the
public of  technical  developments which  nay form  the basis  for  a final EPA
decision, position,  or regulatory action.

                  Standards Development and Support Branch
                    Emission Control Technology Division
                Office of  Mobile Source Air Pollution Control
                    U.S. Evnironmental Protection  Agency

-------
                              Table of Contents
I.  Introduction
II.  Discussion of Specifications

     A.  Methodology                                                    6
     B.  Comparison to Other Specifications                             6
     C.  Costs                                                         10
III.   How to Use the Specifications

     A.  Overview                                                      13
     B.  Definitions and Abbreviations                                 14
     C.  Analyzer Technology                                           18
IV.  I/M Analysis System Specifications

     A.  Gases                                                         IP
     B.  Gas Cylinders                                                 21
     C.  Durability Criteria                                           22
     D.  Design Requirements                                           23
     E.  Analyzer Performance                                          30
     F.  Sample System Performance                                     33
     G.  Operating Environment                                         34
     W.  Fail-Safe Features                  -                         35
     I.  System Correlation to Laboratory Analyzers                    42
     J.  Manuals                                                       43
V.  Test Procedures                                                    45

     A.  Traceability of Analytical Gases                              46
     B.  Gas Cylinder Specifications                                   47
     C.  Durability Test Procedures
          1.  Vibration and Shock                                      48
          2.  Sample Line Crush                                        50
          3.  Sample Handling Temperature Effect                       51
          4.  Filter Check and Hang-up                                 53
     D.  Design Requirement Inspection and Test Procedures
          1.  Useful Life                                              56
          2.  Name Plate                                               56
          3.  Sample System                                        "    56
          4.  Sample Pump                                              56
          5.  Sample Probe                                             56
          6.  Sample Line                                              56

-------
                   Table of Contents (continued)


                                                                 Page

D.  Design Requirement Inspection and Test Procedures
    (continued)
     7.  Analyzer Spanning System                                 57
     8.  Analyzer Ranges                                          59
     9.  System Grounding                                         59
     10. System Vents                                             59
E.  Analyzer Performance Test Procedures
     1.  Calibration Curve                                        60
     2.  Resolution                                               62
     3.  Compensation
          a)  Altitude                                            63
          b)  Pressure and Temperature                            66
          c)  Non-Compensated Systems                             69
     4.  Zero and Span Drift                                      70
     5.  Span Drift (see E.4.)                                    73
     6.  Noise (see E.8.)                                         73
     7.  Sample Cell Temperature                                  74
     8.  Gaseous Interference and Noise                           76
     9.  Electrical Interference                                  79
     10. Propane to Hexane Conversion Factor                      83
F.  Sample System Test Procedure
     1.  Sample Cell Pressure Variation, Low Flow and             86
         Response Time
     2.  (See F.I.)                                               90
     3.  (See F.I.)                                               90
     4.  (See F.I.)                     '                          90
     5.  System Leakage                                           91
G.  Operating Environment Test Procedure                          93
H.  Fail-Safe Systems
     1.  Test Procedures for All Systems
          a)  Warm-up Lock-out                                    95
          b)  Low Flow                                            97
          c)  Leak Check                                          97
     2.  Test Procedures for Decentralized Systems
          a)  Automatic zero/span                                 98
          b)  Automatic Read                                      98
          c)  Printer                                             98
          d)  Vehicle Diagnosis                                   98
          e)  Anti-Tampering                                      98
I.  System Correlation Test Procedures
     1.  NDIR Correlation                                         99
     2.  FID Correlation                                      .   103

-------
I.   Introduction

It  is doubtful  that  anyone would  disagree with the  statement  that automo-
biles  (and  trucks) are  part  of  the  air  pollution problem in the country.  In
the  past.  Congress  has  specified, through  the Clean  Air Act  and various
amendments  to  the Act  that  new cars must  meet  certain  emission performance
standards  prior  to  introduction  of  those vehicles  into  commerce.   These
performance  standards  have always been  checked  under closely  controlled
laboratory  conditions  with  sophisticated  equipment.  The  required accuracy
and  performance  of  this  equipment has  always  been specified within  the
context of  the  laboratory  situation.

Recently, there has been more emphasis  on checking  the performance of in-use
vehicles.   This  is occurring through  the implementation of state inspection
and  maintenance programs as  well  as the forthcoming emission repair warranty
regulations  (207(b)j authorized by  the  Clean Air Act.   Further,  the new 1984
Heavy-Duty  (HD)  Truck  Federal  Regulations  specify  an idle  standard as well
as  a driving  cycle  standard.   Both the  207(b) warranty emissions  test for
hydrocarbons  (HC) and carbon monoxide  (CO), and the Heavy-Duty idle emission
test  for  CO are expected to  be  conducted  on in-use  vehicles with data gener-
ated  mostly by  state-run I/v programs.

The  economic  implications  about  the   accuracy  of  the  data from  these I/M
programs  is of concern.   The 207(b) emission repair warrantv provisions and
the  Heavy-Duty  idle  test data may be used to determine who  pays for emission
repair  maintenance  on  failed  vehicles  - the  individual  consumer,  or the
vehicle manufacturer.   Additionally, the  Heavy-Duty idle data may serve as a
basis  for  recall  action   since  failure of  the  wn  idle  test  is  the same as
failure  of  any other Federal emission  standard.  For those  failed vehicles
not  covered by  207(b)  or  HD  standards,   it  is possible that  the  amount of
repair  maintenance performed will  have  some relationship  to  the  degree of
failure - (i.e. only  those repairs  that are needed  to bring the vehicle into
compliance  will be performed).   If this  minimal maintenance  scenario comes
to  pass,  then the accuracy  and variability of  the  measurements system will
play  an  important  role  in  the dollar amount of the  maintenance sold.

Because   the  I/M  emission check (as  well  as  207(b)  and  HD idle)  is  less
sophisticated  than the  Federal driving cycle compliance checks, it might be
expected  that  the requirements necessary to achieve acceptable accuracy and
^variability with  I/M  analysis systems  would be less sophisticated than those
used  in  the  laboratory.   To some  extend  this  statement is  true,  but it is
not   totally  true.   Consider   for   instance,  the  laboratory analyzers  are
operated  by  trained  and  skilled   technicians.   These  personnel   can  spot
problems  in analysis sometimes even before they happen.  Complete engineer-
ing   departments  are   constantly  checking  the  laboratory  emission  values
against  past values  and   design  goals.  Under  these  conditions,  errors  in
measurement  can be detected and  eliminated.   In a  sense,  the human- mind is
adding  considerable  sophistication  to  the  equipment.   While not saying that
I/M  analyzer operators do  not  have the  mental capacity  to comprehend the
variety  of  emission analysis problems,  it can  be  said  that the typical I/M

-------
analyzer operator  generally has not had  the  training  or  experience  that  the
laboratory  technicians and  engineers have  had.   This  training problem  is
compounded  even  further   in decentralized  I/M  programs.   Mechanics  in  a
repair shop  just  cannot be  required  to  understand the gamut  of measurement
problems.  Even  though mechanics must by necessity become  familiar  with  the
I/M  equipment,  there  is   an economic  incentive   to  direct their technical
skill toward repairing vehicles.

Another  significant  difference  between the  laboratory analyzer,  and the  I/M
type analyzer  is  the environmental operating conditions.   Laboratory analy-
zers are never  operated in  hostile environments.   They are generally  oper-
ated in  heated and  air conditioned  buildings  with  humidity  control.    I/M
programs, on  the  other hand, expect analyzers  to  operate from California to
Maine,   from  Texas to  Minnesota,  and  from summer  to  winter in a variety  of
enclosures  ranging  from  rain  and wind  shelters  to permanent  structures.
This  variety  of   hostile  environmental  conditions  places a  much   greater
burden  on  an  I/M  analyzer than  a  laboratory   analyzer  ever  encounters.

The  lack of  measurement savvy by  the  I/M analyzer operator,  and the signi-
ficant variation  in environmental  operating conditions  are just two  among
many reasons that would suggest an I/M analyzer should be more sophisticated
than a  laboratory instrument.   However, the goals  of the I/M  program,  and
the cost of the I/M analyzer  to the user must be  kept  in  mind.

With these ideals  in mind,  the  important parameters  in the  emission  measure-
ment process  were  evaluated.  A balance among  capability,  cost,  complexity,
and  operator  skill has hopefully been achieved in the recommended I/1*  anal-
yzer specifications presented in this document.

-------
II.  Discussion of Specifications

A.  Methodology

To  those  familiar  with I/M  analyzer  specifications publicized  by  other
organizations (State of  California Bureau of Auto Repair,  Equipment  and Tool
Institute,  etc.),  the specifications  included  in  this document will  appear
somewhat  different even  though they parallel the 1980 California  specifica-
tions  in  some  areas.   To  the  user  not  fully acclimated  to the  emission
measurement  process,  the publishment  of yet  another specification that  is
different  from  previous publications  could   cause   confusion  about  which
document  should be  used.  To avoid confusion about the source of  the  speci-
fications  in this document,  a brief  description  of  the  procedure used  to
generate  the  specifications  is presented,  followed by a  description  of  the
significant  differences  between  this  document  and  other  specifications.

Prior  to  assigning   any values  to any measurement  parameters  (accuracy,
drift, etc), a review of all of the parameters  affecting the  I/M measurement
process  was  made.   A model  of error propagation was  then  formed to  show the
interaction  of  the  various  parameters.  Approximate  ranges of errors  were
assigned  to  each  parameter  based on test data, experience, and/or engineer-
ing  judgment.   Previously published specifications  also  provided many  ini-
tial values.  Using  classical  statistical  approaches, this model  could  then
evaluate  the impact  of  each  specification  on  the   accuracy  of  the  final
emission value.

The  important aspect  of the  model is that  it  focuses   on  the  real  issue
involved  in I/M measurements.  That issue is:   What is the acceptable  accur-
acy  of  the  I/M measurement  process?  After the answer  to this question  is
known,  the parameter specifications  can be adjusted to  insure the desired
result.

There is  a wide variety  of opinion about what  is accurate  enough.  A prelim-
inary  evaluation  of  I/M analyzers  with  the   model   indicate  the error  in
current I/M systems could be on the order of +  25 percent  (50 percent  range)
or worse.  EPA laboratory tests have already shown a  25-30 percent variation
between  analyzers,  even  though  the analyzers  were   set up   and operated  by
highly skilled technicians under favorable  environmental conditions.

The staff  at the EPA laboratory in Ann Arbor feels that the  real variability
in the field among I/M analysis systems  is  unknown at  the  current  time.   The
staff also  feels  that the preliminary data of 25 to  50 percent variability
is  too  large to  meet I/M,  207(b), and  HD  goals  while maintaining  consumer
acceptance of the impartiality and fairness of  the I/M program.

B.  Comparison to Other  Specifications

As  stated earlier,  this specification roughly  parallels  the  technical  por-
tions of  the 1980 State  of California  specifications.  No  attempt  is made  to
recommend  accrediting procedures for  analyzers (other than  a brief discus-
sion in  Chapter  III  Section A).  However,  there  are  differences  in content
as  well  as  differences  in  techniques  for  specifying the  same  or similar
parameters  between  this document  and the  1980  California  specifications.

-------
These differences  are  the result of applying  the methodology  in  addition  to
the  conceptualization  of  the  actual step by  step  procedures  that could  be
used to verify  compliance to the recommended  specifications.   Some specifi-
cations in other documents, although historically specified, almost defy any
type  of  verification  measurements.   Foremost,  it  was  considered that any
specification that  was  recommended  should be  capable of convenient measure-
ments  (although  possibly repetitious and  dull)  by  any emissions  laboratory
with  typical  test  equipment   (i.e.  digital  voltmeters,  chart   recorders,
analytical gases, etc.).

The  concept  of  parameter verification  lead  to the  inclusion of step  by  step
procedures that  can be  used for verification  by anyone.  No  other specifi-
cation document  provides  such  procedures.  While  some  may claim that  such
procedures are not  necessary or that they reduce flexibility,  experience  in
emission analysis  has  shown that there are  many ways to interpret even  what
appears  to  be  straight  forward specifications  with  strikingly divergent
results.    The  recommended  verification   procedures,  therefore,  provide   a
framework  for  interpreting  the  specifications  in  a  consistent   manner.

Another area of conceptual  differences  in  the  verification process  is the
use  of  statistical  techniques  on  the  verification  data.   The  use  of the
statistical  techniques  are  necessary  to   insure  that the verification  data
actually represents the  true  performance of  the analyzer  under  test.  The
use  and extent  of  statistics is not overly  done, and the tendency is to use
them in the  more important areas (e.g. accuracy)  to reduce the  measurement
burden.

A  final area of  conceptual difference between this  specification and others
is the inclusion of specific quality control  features into the basic analy-
sis  system specifications.  Generally, these  .features  are classified under
the  category of "fail-safe"  sub-systems.   In addition, many  of the other
analysis system  requirements contained  in this  document are specified  in  a
manner that  reinforces  the fail-safe concept.  The  degree  of  sophistication
of the built-in quality control  features  is  determined by the  end use of the
system.   Analyzers  intended  for  use  in   decentralized  systems generally
require more sophisticated features (than centralized  systems),  due  to the
more hostile environment and less sophisticated operators.

The highly desireable aspect of  these built-in features  is  that they  tend  to
improve the  overall credibility of  the I/M  program.  The improved credibil-
ity  is obtained  because the analysis system requires the operator to comply
with  good  measurement  practices as well as  preventing  the  operator  from
committing  improper practices.  Both  of   these  actions  improve  the  overall
accuracy of  the I/M results.

A  hidden advantage  that makes  these fail-safe features  even more desireable
is  that  the  forced  improvement of  operator practices  relieves some  of the
burden on  state  audit  teams.   Because  of  these  features, it may  be possible
to perform less frequent audits, and still obtain better quality  I/M  results
than  with  more  frequent  audits of  systems  without the  built-in   checks.

-------
Traversing  from conceptual  differences  to  more  specific  differences,  the
following items represent the more significant areas of difference.

     1.   Analyzer Calibration Curve;   The  analyzer  calibration  curve  re-
     quirements are  more statistically  sound  than other publish specifica-
     tions.   Since  the  calibration  curve  sets  the  basic  accuracy  of  the
     analyzer,  it  is  important  to measure the true performance when attemp-
     ting  to  verify compliance  to  the  specifications.   Additionally  these
     recommendations  require better  accuracy  than other  specifications at
     the projected  207(b)  emission warranty levels of  220  ppmh WC and  1.2%
     CO, and the Heavy-Duty  idle standard  of 0.47% CO.

     2.    Analyzer Fail-Safe Systems;    These   recommended   specifications
     include requirements  for certain lock-out features in  order to prevent
     improper operation of the analyzer.   These fail-safe systems are segre-
     gated  into  systems  applicable to  all analyzers,  and  those systems
     addressing the  unique  problems associated with decentralized  programs.
     The  decentralized   fail-safe  systems will most  likely  require  an  on-
     board microprocessor.

     3.  Analytical  Gases;   All analytical  gases must  be  traceable to  Na-
     tional Bureau  of  Standards,  Standard Reference Materials  (SRMs),  or an
     EPA Office of Mobile Source's approved  standard.  A procedure  is refer-
     enced for determining traceability.

     4.   Analyzer Measurements;    A  policy  decision  was made  that all  I/M
     measurements  would  be  considered  on a  "dry-basis".   This decision is
     consistent with  the  1984  Federal  P.D.  Truck idle standard,  the  pro-
     jected 207(b)  requirements,  and the  effective policy  of  many  state I/M
     programs  (due to the  act  of  using a  refrigeration  water trap).   The
     effect of  this decision would be  to consider  all  current and future
     data  as  dry-basis  data.   However,  the recommendation  to implement  a
     recommended  specification  for  dry  measurement hardware  should be  de-
     layed  until   January  1986.   For  analyzer  manufacturers who  wish to
     implement  dry  measurement  systems  sooner, they  may  do so by  either of
     the  two   techniques permitted  -  electronic compensation  or  physical
     water  removal.   The effect of  implementing  a dry-basis measurement is
     to remove a potential 7-10% variability in measurement.

     ,5.  Analyzer Spanning Concepts;  The  recommended specifications require
     all  analyzers to be  gas spanned  at least  once  a week and electrical
     spanned  at  least  once  an  hour.   More  frequent  gas  span  checks  are
     recommended  every  4 hours.   Two alternatives  to the  4 hour gas  span
     check are  possible  which would  reduce  the  use of analytical  gas.   One
     allows the use  of  sample  cell  pressure  and temperature  compensation.
     The other  allows analyzers with a  basic  accuracy significantly better
     than  that required  in these  specifications to  demonstrate  that  even
     considering  the degradation  of accuracy  (due to  external  effects of
     pressure  and  temperature),   the analyzer  would  still be  within  the
     recommended accuracy specifications.

-------
The  weekly gas  and hourly  electrical spanning  concept is  consistent
with projected requirements  for 207(b) and  the H.D.  idle standard.  The
frequency of gas spanning is less  than emissions  testing experience and
engineering judgment would dictate, but more  frequent  than  other  speci-
fications which generally call for  a 30 day gas span check.

6.   HC Measurements;   Most  I/M specifications  require reporting of HC
as  ppm  hexane.  Since  the   span  gas  is  propane,  and  the  analyzer re-
sponds differently  to  propane  than hexane, a propane  to hexane  factor
is required.  Additional verification of  the  relation  of this factor to
the  composition  of  exhaust  hydrocarbons  is  generally  required by com-
paring the  I/M analyzer performance to the performance  of  a  laboratory
NDIR HC analyzer.

These general  requirements  are included  in this  document but with more
definitive  procedures  for   developing  the factor  and  conducting the
comparison.   But,   there  is a more  fundamental  question  involved in
determining HC measurement criteria.  The point is  that  the Non-disper-
sive Infrared  (NDIR)  HC analyzer  is  limited  in its ability  to measure
all of the hydrocarbons in the exhaust.   Current  NDIR  analyzers tend to
measure  only  straight chain paraffinic_ hydrocarbons,  although they do
have some response  to aromatics and aldehydes.  The  lack of response to
these non-paraffinic  compounds is the primary  reason why the Federal
Test procedure  changed  from NDIR  HC measurement  to FID (flame ioniza-
tion detector) HC measurement  8 years ago  (1972).

Even  though  catalyst vehicles  tend to  produce more  paraffinic  hydro-
carbons   than   non-catalyst   vehicles,   chromatographic   analysis  of
gasoline-fueled engine  exhaust indicates significant  quantities  of the
non-paraffinic  hydrocarbons  even  in  catalyst vehicles.   The  very same
compounds  that  current  NDIR HC analyzer-s have difficulty in  measuring.

Two  questions  arise  from  this discussion -  "Why measure  HC   at  all if
the  only measurement technique  is NDIR?",  and  "Why  can't  the  FID be
used in  I/M work?".  First, although  the NDIR does  not  respond well to
all hydrocarbons, a reduction  in hydrocarbons  that  it  can measure still
tends to reduce the overall  HC output  from  a  vehicle.  Further, in many
cases  reduction of paraffinic HC can  also  reduce  other  classes  of
hydrocarbons.

The  answer  to  the   second question is  that  an FID is a complex piece of
laboratory equipment that requires  a mixture  of hydrogen and  helium for
fuel, and  purified air  for an oxidizer.   At this point,   the staff is
somewhat  skeptical  about the  compatability  of  an  FID  in  an I/M test
environment.

The  response  characteristics of current  NDIR analyzers  could, however,
be  improved,  and new  less   complex techniques could be developed.  In
order  to  provide   an incentive  for  this  advancement  in technology,
beginning  in  January  1986,  it is  recommended that all I/M  analyzers
correlate  to  an FID HC analyzer by the procedures provided.   To  assist

-------
                                     10
     in the degree of correlation that  is  technically  feasible,  it is  recom-
     mended during  the  verification of current analyzer  performance  that  an
     additional  test  be performed  that compares__the  I/M HC analyzer  to^ an
    _FID.   Review of the NDIR/FID correlation datageneTaTe3™B7™th>e—1accredi-
     tation  procedures   is  recommended prior  to  the  implementation of  the
     1986 specifications.

     7.  System  Leak Checks;   Leaks  in the sample  system are  probably  the
     source  of  the largest  and most  frequent errors  that occur  in  practi-
     cally  all   emission  measurement  systems.   This  is because  a leak  is
     transformed  directly  into an  error  (i.e.  a 15%  leak is a 15%  error).
     Most  laboratories  have rigid procedures for  leak checking of  analysis
     systems, and  the  process of  searching  for a  leak can  be  very  time
     consuming.

     I/M analysis  systems  have special problems that  the laboratory  systems
     do not.  First,  laboratory systems are generally not moved around;  I/M
     systems  in  decentralized  systems  are.   Two,  laboratories  have  special
     equipment to  identify  leaks,  the expertise to  use the equipment,  and
     the knowledge to  repair  leaks   inside  the  systems;  the   I/M  operator
     generally does not have  this knowledge  available.   Three, the  labora-
     tories generally have a strong committment to prevent and  repair  leaks,
     and subsequently provide  resources of time and  money for  this  commit-
     ment;   the  independent decentralized  system operator may not have  this
     commitment,   and  may not  be  able  to apply the  resources to  it.   Four,
     the flow rates used in I/M systems are so small that even  a 10%  leak is
     difficult to  measure  without laboratory equipment,  let  alone a  2  or 3%
     leak.

     Most other  specifications do not  provide for a routine  leak  check,  and
     if they  do  provide  for equipment, the equipment  is usually a  tapered
     tube flow meter.   A tapered  tube  flow meter is not  practical for  field
     use.    They  are a  high  maintenance item if built  into  the system,  and
     tend  to  stick from  a combination of  hydrocarbons  and  water.  This  is
     true even  in a  laboratory situation  with  overkill on filter  changing
     frequency.    If the flow meter is  used  to  monitor system  response  time
     as well,  the capacity  of the  flow  meter is  too  large  to  read  leaks
     without  a 15-20% error in the reading.   Primarily  for  these reasons a
     flowing span gas leak check  (through  the sample line) is recommended  on
     a weekly basis.

C.  Costs

In all  likelihood,  the  additional features and specifications  will  increase
the retail  cost   of  the I/M analysis  system.   Developing incremental  costs
for each change  in specification or  additional  feature may not be  possible
(at least not without  extensive analysis)  due  to  the degree of interaction
between the  features.   Another aspect  of  cost, however,  is  the  overall  cost
to the  economy.   More  precise  and fail-safe analyzers may well  decrease  the
cost of performance checks on  the analyzers conducted by state  audit  teams.
Furthermore,  more  accurate  and less   variable systems  will  tend  to  improve

-------
                                     11
allocation of  the  repair costs and  the  need for maintenance  to  those  vehi-
cles  that really  need  it.   To  the  consumer  with  the marginally  passing
vehicle,  that  could  get  by with  less  maintenance because  of  the better
discrimination, it would also be a cost  savings.   Further  the  more  accurate
fail-safe analyzer makes it considerably more  difficult for an unscrupulous
operator  to  defraud  a consumer,  or  for a well meaning operator  to make  an
error  in  analysis.   The prevention of  fraud  or  accidental errors  may  also
lead to cost sayings.

When costs are  considered not only  the  incremental cost of the  analyzer  to
the user should be considered, but the total cost and  impact on society must
be considered as well.   The more  accurate system has  the potential  for  a net
savings to the  national  economy  plus having positive environmental  benefit.

Even though  the benefits to society do  exist,  the  cost of  the  better system
to the system user must  be reasonable.   The  EPA staff estimates that approx-
imately 50,000 pass/fail I/M units will  be sold to  support  the  I/M  programs.
Assuming 10 analyzer manufacturers _!_/ actively market this  type of  analyzer,
and assuming equal market share, each manufacturer would be able  to spread
the development costs across approximately 5000 analyzers.

Development and modification costs could be as  high as  5100,000 2f on some
systems.   Spread  across 5000  analyzers this  would  be $20  per  analyzer.
Using a 3  to  1 mark-up  which  is  standard industry  practice _2_/,  would  cause
the  retail  price  to increase  $60 per  analyzer  due  solely  to  development
costs.   Since  current analyzers  are in  the $2500-$3500 price range,  it  is
apparent that development costs are not  an overwhelming burden.

A list of assumed system changes  is provided in Table  1 along with  estimated
retail  cost   increase.    These  costs  represe.nt an average  estimate,   some
systems may require more improvement, while  others  may  be less  costly to im-
prove.    The  basic   group   of  improvements  ($690)   are  designed  for  well-
trained,  competent  analyzer operators.   These  improvements are  expected  to
reduce current measurement  errors of  25  to  50 percent  down to the  10-  to  15
percent range.  The  cost for these improvements are  estimated  at $17 to $69
for  each  percent  of improvement  in  measurement accuracy  and variability.

For decentralized programs,  accuracy and repeatability of  emission  measure-
ments by  less  sophisticated  analyzer operators  can be  compensated for  by
adding  a  microprocessor  to  the analysis system.   The  microprocessor allows
simpler  external   system  controls  and   automatic  operation.   The  simpler
operation does  increase  the  cost of  the system.   However, considering the
cost to  the  consumer,  and the  great  potential to inhibit improper system
operation and  fraud,  the increase in  cost  is  reasonable and well  worth it.

\j Eight(8) Manufacturers participated in the December  20,  1979 Equipment
   and Tool Institute development of performance test specifications for
   garage type analyzers.

2j Based on informal conversations with  industry representatives.

-------
                                    12
                                   Table 1
           Item                                   Retail Cost Increase
1.   Basic Improvements

     a)  Development                                       $ 60
     b)  Gas Spanning System                              5150
     c)  Improved Detector                                $210
     d)  Improved Signal Conditioning                     $ 80
     e)  Leak Check System                                $120
     f)  Sample Cell  Heater                               $ 70
                                                         $690
2.   Decentralized System Safe Guards

     a)  Microprocessor Based System                       700
                                                         $1390
    Options

     a)  Anti-dilution                                    $600
     b)  Printer                                          $250
     c)  Data Collection and Storage                      $300

-------
                                     13
III.  How to Use the Specifications

A.  Overview

As  they  are stated,  the analyzer specifications  listed  in Chapter TV have
meaning only when  properly defined.   The test procedures presented  in Chap-
ter V provide this context.  These test procedures are  intended  to provide a
consistent  technique  to verify analyzer performance.   As  such,  these tests
should only be conducted  by a  laboratory  with  sufficient facilities,  and
personnel experienced in emission measurement.

The test procedures should be performed in the following order.

     1.  Follow manufacturer's initial start-up and pre-test procedure as
     listed in the manufacturer's manual(s).

     2.  Durability Tests

          1st - Vibration and Shock Test

          2nd - Sample Line Crush Test

          3rd - Temperature Effect

          4th - Filter Check

     3.  Inspect Design Requirements

     4.  Analyzer Performance

     5.  Sample System Performance

     6.  Operating Environment Test

     7.  Fail-Safe Features

     8.  Correlation Tests

After  the analyzer  passes these tests, the  user  can be fairly  confident of
the in-use  emission  test results.   However,  the  confidence applies only to
the analyzer tested.

As  stated  before,  this  document is  not  an  accreditation  procedure.  This
document presents  specifications  and procedures  for determining the  perfor-
mance of an individual analyzer, not a product line.

In  order  to  develop  confidence  in  the  manufacturer's  product  line  as a
whole, additional analyzers must be  tested.   The  number of  additional analy-
zers  tested determine  the  confidence level  of  the product  line.  For in-
stance, every  single  laboratory analyzer used at the  EPA  Laboratory in Ann
Arbor, and at most auto manufacturers' facilities are thoroughly checked out
before they are put  into service.   In many  instances, the analyzer is re-
quired to meet  performance checks even before it is paid for.  This  measure
is probably impractical for the large volume  of I/M instruments.

-------
                                    14
One  practical  solution would  be to  perform a  selective  enforcement  audit
(SEA) as  part of  the analyzer  manufacturer's  QA/QC  procedures.   EPA  uses
this  SEA  principle to  evaluate emission  compliance of automotive manufac-
turers'  product lines.  The SEA sample plan  is a sequential sample plan, and
relies  on  separate criteria  (number of  failed  units)  to  determine a  pass
decision versus  a fail decision.   Such a sample  plan can  determine with  a
reasonable  level  of  confidence  compliance  of  a product from test data  on
less than 1 percent of the product line.

Sample  selection,  procedures,  and  test plans can  be found  in 40CFR 86  Sub-
part K  and  Appendix X (both in the Federal  Register,  January  21,  1980).   An
inspection  plan   incorporating  an  audit  quality  level (AOL) of  10% would
insure  with 95 percent  confidence level that no more  than  10  percent of the
production would fail to comply with  the specifications.

A sampling  plan  for  determining product line compliance (over 400 units  of
production) is presented  in Table 2.  If the market share  is  5000 I/M units
per analyzer manufacturer, as suggested by the cost section, the upper  limit
of 50 units  to  be tested represents  only 1  percent of  production.  In  auto-
motive  SEA  testing,  the  determination of  pass/fail  is usually determined
with much  less  testing  - normally around 12 to  17 units.   Twelve I/M analy-
sis  systems  tested would  represent  only  0.24  percent  of  production which
should not be a burden to the manufacturer.

There are  other  types  of  production  quality   checks  that could  be  used.
However, the advantages of the SEA type check of a small sample  size as  well
as  a sequential  nature  may  outweight the  possible  improvements  of other
checks.   In fact, if  tigher quality is required, the SEA sampling plan  could
be adjusted for  yearly  production rates  (see 40CFR 86, Appendix X, Federal
Register, January 21, 1980).

Once  the  analysis  systems  are delivered,  periodic  performance  checks  of
certain  variables can  insure  proper  operation.   However,  these periodic
checks  are  useless  if the  analyzer owner  cannot  obtain  effective repair
service and parts.  Further, performance checks  after  maintenance are neces-
sary to insure that the analyzer was  repaired properly.

The  previously mentioned  items are beyond the  scope of this  report, but  it
was  felt that  they should be mentioned in order to make the reader aware  of
some of  the  additional  factors involved in  obtaining  reasonable measurement
results.

B.  Definitions and Abbreviations

The  following definitions are somewhat general  but  are provided for refer-
ence.   For  many  definitions,  the  specific  test procedures used for verti-
fication of the specifications provide  the final interpretation.

-------
                                     15
                      10% AOL
    Table 2

SEA Sample Plan
 Over 400 Units Produced
No. of
Failed Units
For
Pass Decision
(1)
(1)
(1)
(1)
(1)
(1)
0
0
0
0
0
1
1
1
1
1
2
2
2
2
2
3
3
3
3
3
4
4
4
4
No. of
Failed Units
For
Fail Decision
(2)
(2)
3
3
3
4
4
4
4
4
5
5
5
5
5
6
6
6
6
6
7
7
7
7
7
8
8
8
8
8

No. of
Units
Tested
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
46
47
48
49
50
No. of
Failed Units
For
Pass Decision
4
5
5
5
5
6
6
6
6
6
7
7
7
7
7
8
8
8
8
8
9
9
9
9
9
10
10
10
10
12
No. of
Failed Units
For
Fail Decision
9
q
9
9
9
10
10
10
10
11
11
11
11
11
12
12
12
12
12
13
13
13
13
13
13
13
13
13
13
13
No. of
Units
Tested

  1
  2
  3
  4
  5
  6
  7
  8
  9
 10
 11
 12
 13
 14
 15
 16
 17
 18
 19
 20
 21
 22
 23
 24
 25
 26
 27
 28
 29
 30
(1)  Test sample passing not permitted  at  this  stage.

(2)  Test sample failure not permitted  at  this  stage.

-------
                                16
1.   Accreditation:   Certification  that  an  analysis  system  has been
tested by  the required  procedures,  and it has  been verified that the
analysis system  complies  with those specifications.  Further, accredi-
tation implies that the analyzer manufacturer follows proper procedures
to  insure  that  subsequent  production systems  are also in compliance.

2.  Accuracy;  The combination of bias and precision errors that quan-
tify  the  difference between  the analyzer  reading and  the  true value
(see equivalency).

3.  Analytical Gases:  Gases  of known concentration used in the  analy-
tical process as a reference.  The three basic categories are:

     a) Calibration Gas
     b) Span Gas
     c) Zero Gas

4.   Analysis System;   A  system  that  incorporates  an  analyzer(s)  and
sampling components for the purpose of exhaust gas analysis.

5.  Analyzer:   A  device  that has  the  capability  to  identify unknown
concentrations of  particular  constituents  in automobile or  truck ex-
haust  gases  by  comparison  to analytical  gases.   Commonly  used  inter-
changeably with  "instrument".

6.  Calibration Gases:  Analytical gases that are  used to determine the
accuracy of an analyzer calibration curve.

7.  CO:  carbon monoxide

8.  C0?;   carbon dioxide

9.  Detector;  The portion of  the analyzer that detects  the constituent
of interest,  and provides the original signal proportional to the con-
centration of the  constituent.

10.  Drift;   The amount  of change with  time  of analyzer reading.  Two
components of drift are:

     a) Zero drift - change in zero reading
     b) Span drift - change in the difference between zero and span
                     readings

11.   Dry-Basis  Concentration;   The  resultant  concentration  after the
water  has  been  removed  from the sample either  physically  or by elec-
tronic simulation  techniques.

12.   Equivalency;   A  statistical  comparison of  a  candidate analysis
system performance versus the reference analysis  system performance on
exhaust gas  in  order  to determine the  acceptability  of the candidate
system.

-------
                                17
13.  FID;   Flame  ionization detector.  Most  common  laboratory  analyzer
used for  determination  of  hydrocarbon (HC)  concentrations in exhaust
samples.

14.  fs;  Full scale of  the analyzer.

15.  Hang-up;   Hang-up refers to  the  process of hydrocarbon molecules
being  absorbed,   adsorbed,  condensed,  or  by any  other  method removed
from the  sample  flow prior to reaching the analyzer detector.  It.also
refers  to any subsequent  desorbtion  of  the  molecules into the  sample
flow when they are assumed to be absent.

16.  HC;  hydrocarbons

17.  Instrument;  see analyzer

18.   Interference (electronic);    Analyzer  read-out  errors  caused by
instrument  response  to electromagnetic sources  and power  supply  varia-
tions.   Common forms of  electromagnetic sources  are:

     a)  Radio Frequency Interference  (RFI)

     b)  Very High Frequency Interference  (VHF)

19.  Interference  (gases);  Analyzer read-out errors  caused by instru-
ment  response  to  non-interest  gases typically  occuring  in vehicle
exhaust.

20.  L.S.;  Low scale or range of  the  analyzer.

21.  NDIP;  Non-dispersive Infrared Analyzers

22.  Optical Bench;  The portion of the  analyzer  that consists  of  the
main sample  processing and detecting  assembly.  Generally included  are
the  detector,  optical  filters,   sample   tubes,  infra  red source,  and
chopper systems.

23.   Precision;   Statistical  quantification   of   random measurement
errors.

24.  ppm;  parts per million by volume

25.  ppm C;  ppm by Carbon atom

26.  ppm C3 or ppmp;  ppm propane  (C   H )
     J_L______^___t_l_    •              j  J^

27.  ppm C6 or ppmh:  ppm n-hexane

28.  Propane  to Hexane Conversion  Factor;  A  factor that accounts  for
the difference in analyzer response  (relative response) between propane
and  n-hexane.   Sometimes  referred  to as a  "C" factor,  or  a propane
equivalence factor  (P.E.F.).

-------
     29.  Response;  Analyzer indication to a gas.

     30.  Response Time;   The  reaction time between  a  change in concentra-
     tion at  the  inlet to the sample system and  the  time the analyzer indi-
     cates a given percentage of that change.

     31.  Sample System;  The portion of the analysis system  that is respon-
     sible for delivering an unaltered sample to  the analyzer.

     32.  Span Gases;   Analytical gases  that  are used  to  adjust or return
     the analyzer response  characteristics to  those determined by the cali-
     bration gases.

     33.  Zero Gas:    An  analytical   gas  that  is  used  to  set  the analyzer
     response at zero.
C.  Analyzer Technology

These specifications were  determined based on current I/M practice of using
non-dispersive  infrared  (NDIR)  analyzers  for  HC  and  CO  measurements.
Nothing  in these  specifications  should  be  contrued as  prohibiting other
analysis techniques.   Potential  improvements  in technology should be consi-
dered on  a case  by case basis.   To that extent, many  of  the concepts ex-
pressed by these specifications can be applied.

-------
IV.  I/M Analysis System Specifications

A.  Gases

     1.  Accuracy

          a) Calibration gases:  + 2% of True Value
                                 + 1% of NBS Standard Reference
                                 " Material  (SUM)

          b) Span gases:  + 3% of True Value
                          + 2% of NBS Standard Reference Material

          c) Zero gases:  less than 10 ppm C hydrocarbon
                          less than 50 ppm CO
                          less than 1000 ppm CO

     2.  Composition

          a) CO and HC  (propane) with N  or air diluent

          b) CO  with N  or air diluent

          c) Zero  gas  may  be bottled  gas  or  chemically  purified room  air
          such  as  with an  activated  charcoal  trap  on  the analysis  system.

          d)   Fexane  to   propane  conversion   factor  checking  gases  shall
          consist  of  n-hexane in  nitrogen,  gravimetrically  blended  to  the
          accuracy of  SRM's.   The  gravimetric  analysis  is only valid  for  1
          year  from  the date of analysis unless  historical data and corre-
          lation checks can verify stability erf  the gas concentration.   The
          gravimetric analysis is void if the gas cylinder temperature  drops
          below  20°C  (68°F) for any  reason  including  shipping and storage.
          The required concentrations are:

                i)  200 ppmh  ( +_ 15 ppmh)
                ii) 1800 ppmh  ( +_ 150 ppmh)

     3.  Recommended number of gases

          a) for new instrument check out

                i) 7 equally spaced concentration  values  of calibration
                gases per range

                ii) bottled  zero gas

                iii) 2 concentrations of hexane/propane conversion  factor
                checking gas

-------
                          20
b) for in-use systems

     i) one span gas concentration between 70-95% of fs on  low
     range for each HC and CO analyzer

     ii) room air zero gas

c) for periodic check (i.e. state inspection) of in-use systems

     i) a minimum of 3 concentrations per range for each analyzer,
     or

     ii) one concentration for each cutpoint with a concentration
     value within 10 percent of the cutpoint (need not exceed 5
     concentration levels), and

     iii) bottled zero gas

-------
                                    21
B.  Gas cylinder specifications

     1.  Span gases

          a)  Gas cylinder should be of a 1A size.

          b)  All cylinders shall meet DOT specifications for 1A cylinders.

     2.  Concentration Label

     A semi-permanent label shall be affixed or attached by the Certifying
     Laboratory to each gas cylinder with the following information.

          a)  Name of the Gas Blender

          b)  Name of the Laboratory Analyzing the Gas Blend

          c)  Cylinder I.ID. number

          d)  Date of Analysis

          e)  Traceability to NBS or to other certified EPA Mobile Source
             Standard

          f)  Statement of Impurities

          g)  Gas Concentration

-------
                                    22
C.  Durability Criteria

     1.  Vibration and Shock Test
         (See Test Procedure)
     2.  Sample Line (see Test Procedure)

          a) Line crush
     3.  Sample Handling System Temperature
         Effect (20 minutes (? 600°F sample
         inlet, see Test Procedure)
     4.  Filter Check (see Test Procedure)

          a) 2 hr sample time
          b) Sample until "low flow"
              indication
a) zero shift:  2% fs LS
b) span shift:  2% fs LS
c) All analyzer and system
   performance checks must
   be met.
i) No visible failure or
   deterioration
ii) Meet leak check
    specifications
iii) Meet response time
     specifications

a) No visible failure
   or deterioration.
b) Meet leak check
   specification.
c) Meet response time
   specification.
d) Meet HC hang-up
   specification.
i) Low Flow Not Activated
ii) Meet Leak Check
    specifications
iii) Meet Response Time
     Check
iv) Meet HC hang-up check

i) Time until "low flow"
   system activates
ii) Until the low flow
    system activates the
    system response time
    must be met.

-------
                                    23
D.  Design Requirements

     1.  The  analysis system  shall be  designed  for a minimum  of  a 5 year
     useful life.  Useful  life has terminated when  the  analysis system can
     no longer be  repaired to meet the specifications in this document at a
     cost  of  less  than  60 percent of  the replacement cost  at  the time of
     repair.

     2.  NAME PLATE Permanently Located and Readable.

          a) Analyzer System Manufacturer

               i) Name
               ii)  Address
               iii) Phone number (customer service)

          b) Analyzer Detector or Optical Bench Manufacturer

               i) If the same manufacturer as the analyzer manu-
               facturer item b) may be deleted.

               ii)  Name

               iii) Addresss

               iv)  Phone number (customer service)

          c) Analyzer Model Number and Serial Number

          d) Detector Model Number and Serial Number

          e) Date of Analyzer System Assembly

     3.  Sample System

          a)  Sample  system components  should be  designed  for  intermittent
          sampling  (20 minutes  out of  10 hr.)  of  600°F inlet (to probe)
          exhaust  temperature.   Such  systems shall be designated "Unloaded"
          analysis  systems.    Systems  designed  for  continuous  duty with a
          minimum of 1000°F inlet  (to probe) exhaust temperature, and proper
          sample handling equipment (i.e. water removal and/or heated lines)
          may be called "Loaded" analysis systems.

          b)  The  type of  system  (Loaded or  Unloaded)  shall be permanently
          attached and prominently displayed in large letters.

          c) The materials used in the sample handing system  shall. not alter
          the  exhaust  sample.  Some examples  of  non-reactive materials are
          teflon ,  viton~,  stainless steel,  silicone rubber  (red), and in
          some  areas  nylon .   Soie  examples  of  reactive  materials  are,
          brass, copper, and tygon".

          d) Water Trap:

-------
                               24
          i) A water  trap shall be included  in  the sample  system.  The
          trap shall  be  self draining,  visible to  the  operator, and
          shall prevent condensable  water from  occurring in  the  sample
          system downstream of the water  trap.

          ii)  Alternative 1:   Electronic correction  to  a "dry-basis"
          reading  is  permitted if the  temperature  of the sample  after
          .the  water  trap   is  measured and the  sample is  assumed  to be
          saturated at  the measured  temperature.  The water  correction
          system must be deactivated  for  leak checks.

          iii) Alternative  2:   A dry-basis measurement is  permitted if
          the  absolute  moisture content  entering  the analyzer is  less
          than the saturation moisture content  at a  sample gas temper-
          ature  of 7°C  (45°F).   Ice  traps  or  refrigerators are  per-
          mitted.

     e)  Particulate Filter

          i)   A  particulate  filter  shall be  included  in the  sample
          system.

          ii)  The filter  shall be   located  in a manner  that  allows
          convenient filter element replacement.

          iii)  The proper direction  of flow  for the  filter body(s) and
          element(s)   (if   applicable)  shall  be indicated  in a  manner
          that is easily discernible.

          iv)  The filter body shall  be-  designed  to  provide  leak  free
          operation with  normal filter element changing  frequency for
          the  lifetime  of  the  analyzer,   or  a  parts  list  of  specific
          replaceable parts  shall be  listed  in the maintenance  manual
          as parts that may contribute  to  sample system leaks.

          v)   All  manuals and  filter element replacement  instructions
          shall  indicate  that the system  should be  leak  checked  every
          time a  filter element is changed.  A similar message  should
          be located on or near the filter body(s).

          vi)  The filter element and  filter  system  shall be  designed
          to  prevent  particulates larger in size  than 5 microns  from
          entering  the  analyzer.   Verification of  filter particulate
          size removal  is  permitted  to   be  determined  by the  filter
          manufacturer  using  standardized  ASTM  or  Filter   Industry
          procedures.

4.   Sample Pump

     The sample pump shall be designed  for at least a 2000  hour life of
     continuous duty.

-------
                               25
5.  Sample Probe

     a) The  probe must sample  at  least 16 inches  from  the end of  the
     tailpipe or dilution adaptor.

     b) The  probe shall  have a  flexible  portion  that  will allow  the
     probe  to  be  inserted  in  a  1M  inch  O.D.  tailpipe.   Assume  the
     tailpipe has  a  3  inch radius 90° bend beginning 4  inches  from  the
     end of the tailpipe.

     c) The probe shall be designed  for easy servicing and/  or  replace-
     ment.

     d) A universal  leak proof  dilution adapter  (tailpipe extender)
     shall be provided.

     e) The  tailpipe  extender  shall  be  able  to  be attached to  the
     vehicle within 60 seconds.

     f) The tailpipe extender shall  be designed  to  allow  the attachment
     of standard  service  center  building exhaust  evacuation  systems.

     g) The tailpipe extender shall  not alter  the sample  and the mater-
     ial shall conform to D.3.C.) of this chapter.

     h) The probe  and  tailpipe extender shall have sufficient  hardware
     (insulated handles,   etc.)  that  will  allow  the user to insert,
     attach, or  remove the  probe or  the  dilution adapter safely  and
     conveniently.

     i) The  probe or  tailpipe extender shall be  designed in  a manner
     that  will  prevent the  probe or  extender from being removed from
     the vehicle unintentionally.

6.  Sample Line

     a) The  sample  line  shall be  flexible  at  the temperatures  to be
     encountered during vehicle testing (See Section G)).

     b) The  sample  line  shall  not be longer  than  35 feet  nor shorter
     than  10 feet (excluding  the probe).

     c) The  portion  of  the  sample line  in  contact  with the exhaust
     gases shall  not alter  the exhaust sample,  and  shall minimize HC
     hangup,  due  to absorption,  adsorption,  desorption,  outgassing,
     etc.

     d) The  manufacturer  must  state  the estimated useful  life of  the
     sample line in the owners manual  or on the  name plate.

-------
                               26
7.   Analyzer Spanning System

     a)  Overview and Frequency:  All analyzers shall follow the follow-
     ing spanning frequencies:

          i) Weekly - gas span
          ii) Daily

               1) Recommended Basic System - gas span every 4 hours,
                                             electric span every hour

               2) Alternative 1 - electric span every hour
                  (pressure and
                   temperature
                   compensation)

               3) Alternative 2 - electric span every hour
                  (extra basic
                  analyzer accuracy)

     b)  The  analyzer  shall be designed  for  routine gas spanning every
     180 hours (once per week).

     c)   The  gas  spanning operation  must  automatically  correct   for
     changes  between  the  gas  span point  gain  and  the electrical  span
     point  gain.   In  all cases  the electric  span point  gain  is  the
     parameter  that must  be  changed such  that the new  electric  span
     point  gain  setting will  cause the analyzer  to  read  the span  gas
     properly.

     d)  Unless  the  other spanning  alternatives are exercised  the basic
     analyzer shall be designed for routine gas spanning after:

          i) every  "power  on" and warm-up  sequence, and
          ii) every 4 hours of "power on"  condition.

     e)  Alternative Spanning System  1: .  The analysis system may provide
     temperature  and  pressure compensation  to  the analyzer output  and
     spanning  system.   The compensation shall  be  based on sample  cell
     pressure  and inlet  temperature.   If such  a system  is  used,   and
     verified  by subsequent checkout, an  electrical  span check may  be
     substituted  for the "power on" and  4  hour  gas  span checks, but  not
     for  the weekly  gas  span checks.   During analyzer checkout,  only
     one  gas  span  check  is  allowed as  part of  the  initial analyzer
     set-up  prior to  the  initiation of  the entire  check out  procedure.
     An electric span shall be substituted for all gas spanning  opera-
     tions  in  Chapter  V  (unless otherwise  noted).   Additional  tests
     will be  required in order to  verify  the accuracy  and  linearity of
     the  compensation network at  the various atmospheric  pressures and
     temperatures  required by  the performance specifications  (Chapter
     IV,  Secion  E).

-------
                               27
     f)  Alternate Spanning  System  2:   The basic accuracy and precision
     of  the analysis system may be significantly better than the speci-
     fications put  forth  in this document.  In such a case, the analy-
     sis system  may be capable  of meeting  the  specifications  of this
     document without  any pressure and/or temperature compensation of
     the analyzer output and spanning system.  If such a claim is made,
     and verified by  subsequent  checkout,  an electrical span check may
     be  substituted for the "power on" and 4 hour gas span checks, but
     not for  the weekly  gas  span checks.   During  analyzer check out,
     only one gas span  check  is allowed as part of the initial analzer
     set-up prior to  the  initiation of  the  entire  checkout procedure.
     An  electric span  shall be substituted for all gas spanning opera-
     tions   in Chapter  V   (unless  otherwise noted).   Additional tests
     will be  required  in  order to verify  the  accuracy of the analysis
     system  at  the  various   atmospheric   pressures   and  temperatures
     required by the  performance  specifications (Chapter  IV,  Section
     E).

     g)  All  analyzers shall  be  electrically  spanned  every hour after
     "power on" or 4 hour  gas span checks.

     h)  The  analysis  system  shall include  suitable  timers  to insure
     that  the spanning  frequencies are met.   The  timing  systems shall
     also  prevent  use of  the analysis  system by driving  the  readout
     devices  to  full  scale and  prevent any printer  that  may  be used
     from  printing.   Performing  the   appropriate  spanning  operation
     shall   automatically  reset  the  timer for  that   specific  type of
     operation regardless  of  the time  elapsed since that operation was
     last performed.

     i)  The analyzer shall be spanned wi'th flowing zero and span gases.

     j)  The analyzer  system shall  calculate and make  available to the
     operator the span gas set point as ppmh (hexane) based on the span
     gas cylinder concentration in propane, and the analyzer's propane/
     hexane conversion factor.

     k)  The  span point for all  analyzers shall be between  70  and 100
     percent of full scale on the lowest range.

     1)  Appropriate valves,  switches,  and electrical  controls shall be
     provided that will allow the operator to conveniently select zero,
     span,  or sample gases, and zero and span  the analyzer.

     m)  The analysis system shall include a structure  for safely secur-
     ing two 1A size cylinders.

8.  Analyzer Ranges

     a)  Low-Range   :                 i) 0-400 ppmh HC
                                     ii) 0-2% CO

-------
                               28
     b)  High Range  :                i) 0-2000 ppmh HC
                                     ii) 0-10% CO

     c)  An analyzer may use one range for HC or CO corresponding  to  the
     high  range  values  provided  that  the  low range  portion (0-400,
     0-2%) meets all  of  the analyzer specifications as if  that portion
     were a separate  range.   During check out  of  the system  the 0-400
     ppmh  and  0-2% portions  of the high  scale shall  be treated as a
     separate range.

     d)   All  analyzer  read out devices  (digital  or analog)  shall be
     appropriately  scaled  and  capable  of reading negative  values up to
     -10 percent of full scale  for each range regardless  of operational
     mode (sample,  span, leak check, etc.).

     e)   All  HC analyzers  and  manufacturer's  literature  shall indicate
     hydrocarbons  as  ppmh or  ppmC6.   The  symbol  ppm  should  never be
     used for these analyzers.

     f)   All  analyzers shall have  range  indicating  lights  that clearly
     indicate which meter  scale is to be read.  The lights shall be of
     sufficient  intensity and  color  that  will  allow an  operator to
     identify  the  range selection  in  sunlight  from  a distance  of 15
     feet.

     g)   Additional  analyzer  ranges may be used provided  that  they meet
     all specifications pertaining to the recommended ranges.

9.  System Grounding

     a)   All  systems  shall have  a 3  wire power  cord  with a 3  prong
     grounded plug.

     b)  All systems shall:

          i)  have   either  a  ground  fault circuit  that  prevents   the
          analyzer  from  performing analysis functions  (spanning, samp-
          ling, etc.), or

          ii) be tested  with  a 2 wire (non-grounded) extension cord or
          non-grounded  adapter with the power  wires  (hot and neutral)
          in both positions.

     c)   If  a  ground  fault circuit  is  used,   the  system  shall have an
     indicator light  that  activates when a ground fault  is determined.

10.  System Vents

     a)   No  restrictions  such  as flowmeters may be placed downstream of
     any  analyzer   vent  (a  series  analyzer   flow  path  is permitted)
     unless  the system can  detect potential  changes  in restriction
     (i.e. sticking flowmeter), and

-------
                          29
     i) alert  the operator  of the  problem  which would require a
     new  gas  span  and/or  repair  of  the  component  causing  the
     restriction, or

     ii) use automatic compensation  of  the analyzer  readout device
     for the change  in restriction.

b) A  change in  restriction  that  will  cause  a  3 percent of  point
change  in  the  analyzer response  shall  activate the alert system.

-------
                                        30
E.   Analyzer Performance Specifications

     1.   Calibration Curve Uncertainty including Bias errors,  Precision
     errors, and Historesis (Per test procedure) :

          a)  5% of point

          b)  Below 60 ppmh HC or 0.30% CO the uncertainty specification need
          not apply.

          c) If  the analyzer prevents the operator from reading below 10
          percent of full scale on the high scale, then the uncertainty
          specification need not apply to that portion of the  high scale.

     2.   Resolution:

          a) analog meters - 2.5% fs on each range

          b) digital meters

               i) xxxx ppmh

               ii)  xx.xx% CO

     3.   Compensation:

          a) Altitude compensation:   The analyzer  shall have sufficient zero
          and span adjustment to allow the spanning of the analyzer at any
          altitude between 0 and 7000 feet.   The external span adjustment  may
          be limited in range as long as a clearly marked internal adjustment
          will make up the difference.

          b) Compensated Systems:  (Pressure and temperature compensation of
          analyzer read-out and span system)

               i) The temperature compensation network shall provide accurate
               results over the ambient temperature range specified in Section
               G of this chapter as  well as exhaust gas temperatures up to
               55°C
               ii)  The pressure compensation network shall provide accurate
               and  linear results (analyzer read-out)  over a pressure range
               of + 2 inches HgG from the local barometric pressure.   The
               system shall operate between 24 and 31 inches HgA.   Test
               points about which the pressure shall be varied in  order to
               ascertain the accuracy and linearity of the compensation
               network are 24.5, 28.5, and 30.0 inches HgA.

          c)  Non- compensated Systems:  For those analysis systems  that claim
          pressure  and temperature compensation is not necessary,  the analy-
          sis system shall be tested to specification outlined in  step b) for
          pressure  and temperature compensated systems.

-------
                                    31
     4.   Zero  Drift:   + 2%  fs  L.S.  for 1  hour

     5.   Span  Drift:   + 2%  fs  L.S.  for 1  hour

     6.   Noise (clean  environment):   0.50% fs L.S.

     7.   Minimum Sample Cell Temperature:   55°C (131°F)

     8.   Interferences

          a) Gases
               i)  14% CO,
               ii)  Saturated Water
               
-------
                               32
9.   Propane  to Hexane  conversion factor:   The  nean propane to hexane
conversion  factor  for  each  analyzer  sold  as  a  pass/fail inspection
analyzer shall be between 0.48 and 0.56 for each  test point.  The mean
value shall  be known  at the  90% confidence  level  to two significant
figures.  The  confidence interval shall not exceed  a 0.01  increment  in
the  correction factor.   Due to the exceptional hang-up characteristics
of hexane, all components that come in contact with  the hexane gas used
for  determination^ of  the factor shall be  either clean stainless steel
or teflon" (viton  valve seats are permitted).

-------
                                    33
F. Sample System Performance Specifications

     1.   Maximum Sample cell pressure variation
         between gas spanning and sampling   :  4" F20

     2.   Maximum sample cell
         Pressure variation during sampling  :  6" HO

     3.   Maximum sample cell
         Pressure variation between normal
         flow and low flow indication        :  4" HO

     4.   Response Time                       :  12 sec
         Inlet of probe to 95% fs L.S.    ((? low flow indication)
         (See Test Procedure)

     5.   System Leakage Rate                 :  a) sample side 3% of
         (See Test Procedure)                     point on L.S.
                                               b) span gas side, None

     6.   HC Hangup (see Test Procedure)      :  5% fs L.S. in 20 sec.

-------
                                    34
G.  Operating Environment

     1.  The  analysis system  shall  meet  the  analyzer performance  and the
     sample system performance specifications under the following conditions
     (see Test Procedure):

          a) Ambient Temperature:  Between 35°F to 110°F
          b) Relative Humidity:    Between 10% to 99%

     2.  The  analysis  system shall be able  to  be stored at any temperature
     between  -20°F  to  130°F with no adverse effects  on subsequent analysis
     system performance.

-------
                                    35
H.  Fail-Safe Features

     1.  Required Features For All Pass/Fail Systems

          a) Warm-Up

               i)  The analyzer  must have  a warm-up  lock-out  feature with
               indicators.

               ii)  The  lock-out  feature  shall  prevent  operation  of  the
               printer  (if used)  and  read-out meter  until  the  system  is
               warmed up.

               iii) The lock-out feature  shall  be activated  when:

                    1)  the sample  cell is  less than the temperature  speci-
                    fied in Chapter  IV section  E, and

                    2) When  system power is first  turned  on.  The  lock-out
                    shall  stay  on  until  the  zero  drift  is  stabilized.
                    Stabilization  is  determined by observing the  zero drift
                    over a 5  minute period after the  lock-out feature deac-
                    tivates.  The  zero drift during  this 5 minute  period  may
                    not exceed  one-half  of the zero drift  specifications in
                    Chapter IV Section E.

          b) Low-Flow

               i)  The analyzer  must have  a low  sample flow indicator.   If
               the  indicator  is activated,  the analyzer  read-out shall  be
               driven  to  100% of  full scale and the printer (if used)  shall
               be prevented from printing.

               ii) The low flow  indicator shall be activated when  the  sample
               flow rate   is  decreased  to a point  that would not allow  the
               analysis  system  to meet  the response  time  specifications.

               iii) The  low  flow  indicator shall  be prominently  displayed,
               and shall be observable from at  least 15 feet away.

          c) Leak-Check

               i)  The  sampling system shall have an automatic leak  checking
               system for  the sample side of the system.

               ii)  The analyzer  shall  have a timer  that  will  allow  the
               analyzer  to operate  for  180 hours   (once  per week) between
               leak  checks.    If  after  180 hours  the  system  is  not  leak
               checked,  or  the system  fails  a  leak check,  the  analyzer
               readout shall  be  electronically  driven  to 100 percent of full
               scale,  and the  printer  (if used)  shall  be prevented  from
               printing.

-------
                               36
          ill) The  leak  check shall compare  the  analyzer's  response  to
          span gas  with  the analyzer's response  to  that same span gas
          introduced  through  the probe and  sample  line.  The span gas
          flow to  the  probe shall be restricted  such  that the pressure
          in  the  sample  line  during  leak  checking  is approximately
          equal  to  or  slightly  below the  pressure  occurring   during
          sampling.

          .iv) The  system shall  automatically compare the two analyzer
          responses, and make a determination  of  pass  or fail.

          v) A leak check pass or  fail  indicator shall be  prominently
          displayed.

          vi) The analysis  system shall provide a leak tight receptacle
          for the probe on  the system structure for  the  purpose  of  leak
          checking  the system.

2.  Required Features for All Decentralized Pass/Fail  Systems

The  following  fail-safe  features  are  specified in  order  to  prevent
improper operation  of  analysis systems used  for  decentralized systems.
However,  they  could  be  useful  in  centralized  systems as  well.    In
general these features may  require an on-board microprocessor.

     a) Automatic zero/span check

          i) The analyzer shall not have  any adjustments available  to
          the operator  for  adjusting zero or  span point for either the
          gas spanning  operation or the electrical  spanning operation.
          These  adjustments shall  be available in the tamper-proof box
          described in Section H.2.d) of this  chapter.

          ii) The analyzer shall have a selector  switch  or button (with
          indicator light)  labled "gas span".  Activation of the switch
          shall  cause  the  analyzer  to  automatically  perform   the gas
          spanning  sequence with flowing  zero and  span gas  consistent
          with the  requirements for  the manual analyzer  spanning system
          specified in Section D.7.  of this chapter.

          iii) The  analyzer shall have a selector switch or  button  with
          indicator light labeled "electrical  span".   Activation of the
          switch will  cause  the  analyzer to  automatically  perform  an
          electrical  zero and span  operation consistent  with  the re-
          quirements  for  the  analyzer manual  spanning system specified
          in Section D.7. of  this chapter.

          iv) The  automatic  gas  spanning operation  shall   no't  require
          more  than  90  seconds  to complete  once  the "gas spanning"
          switch is activated.

          v) The gas span values shall be  entered via  switches or other
          convenient means  to the following resolution:

               HC = XXXX ppm  propane
               CO = XX.XX%  CO

-------
                          37
     The switches  shall  he in a  tamper  proof hox as described  in
     Section H.2.d) of this chapter.

     vi)  If the  adjusted  span  voltage  changes by  more  than  20
     percent of  the previous  span  voltage,  the analyzer read-out
     shall  be  electronically  driven to  100% of  full  scale, the
     printer  if  used  shall  be  prevented  from  printing,  and  an
     indication  of  analyzer  malfunction  shall  be  prominently
     displayed.

     vii) The automatic checking feature  shall  compute the equiva-
     lent hexane  (HC) span  set  point based  on the input propane
     value,   and  the  analyzer propane/hexane  conversion  factor.
b) Automatic Read System

     i) The  analyzer  shall have a selector switch or button  (with
     indicator light) labeled "Sample" or "Test".

     ii) Activation  of  the "sample" switch shall cause  the analy-
     zer system to begin integrating or averaging the analyzer re-
     sponse  15 seconds after the switch is activated, and continue
     integrating the analyzer response to a flowing sample for the
     next 15 seconds.

     iii) The  analyzer  read-out  device  shall  display  the  inte-
     grated value, and hold the display until reset.  An indicator
     light shall  signal  the operator when the  integrated value  is
     displayed.

     iv) The analyzer shall be prevented  from  printing the  inte-
     grated  value  until  the "sample" switch  is activated  and the
     "sample" cycle is completed.

c) Printer

     i) The  analysis system shall have a printer  that provides the
     consumer a receipt with the following information:

          1) Applicable  cutpoints  or standards  for HC and CO.

          2) Integrated vehicle test values for HC and CO.

          3) A pass or fail indication.

     ii) The print  system shall have appropriate means  (switches,
     etc.) to enter the cutpoints.

     iii) The  system shall determine items  i)2)  and i)3).

-------
                               38
     d) Vehicle Diagnosis

          i)  For  the purpose of vehicle  diagnosis and/or repairs,  the
          analyzer  shall  have a selector  switch  or button (with  indi-
          cator light) labeled "Vehicle Diagnosis"  or "Vehicle Repair".

          ii) Activation  of  the "Vehicle Diagnosis" switch shall  allow
          the  analyzer to  continuously  monitor   the  vehicle exhaust.

          iii) The  printer,   or  any automatic  data collection  system,
          shall be prevented  from operating anytime the analysis system
          is in a "Vehicle Diagnosis" status.

     e) Anti Tampering

          i)  The  anti-tampering feature  shall be  designed  to  prevent
          intentional tampering with the analysis  system.

          ii)  All  switches   or  entry  access  for  automatic  zero/span
          check adjustments,  anti-dilution  limits,  etc.  shall  be con-
          tained  in  a tamper-proof  box or other  tamper-proof mechanism
          with provisions for an inspector's seal.

          iii) The  tamper-proof  system must allow  convenient access by
          an inspector.

3.  Anti Dilution   (Optional  for all systems)

The anti-dilution  feature can perform  three useful functions:   1) most
important, it can serve as an indicator of vehicle  exhaust system  leaks
that would cause the measured emission values  to  be lower; 2) it can be
used to  identify  dilution of the sample  due to either probe placement
or  tailpipe  extender  leaks; and  3)  it  could potentially be  used to
insure  decentralized operators  actually  insert  the  sample  probe  (or
attach  the  extender),  and  actually  measure  vehicle  exhaust samples.

     i) The anti-dilution feature shall identify  vehicle  exhaust system
     leaks and sample dilution.

     ii)  The  preferred technique  for  identifying  leaks  is monitoring
     the CO  levels  in the exhaust.

     iii) At  least  three lower limit  CO. values  shall  be capable of
     being used.

          1) no air  pump
          2) air pump
          3) spare channel

     iv)  The  resolution  of  the CO. span  gas  values entered shall  be,
     XX.X% C0-

-------
     v)  The CO. values shall be entered by switches or other convenient
     means.

     vi) The C0? analyzer shall meet all applicable analyzer design and
     performance specifications  (Chapter IV) between  CO- values of 8%
     and 14%.

     vii) If  the  C0?  is less than the lower limit, the analyzer output
     shall be  electronically  driven  to 100% of full scale, the printer
     (if used)  shall  be  prevented from printing,  and  an indication of
     exhaust system dilution shall be prominently displayed.

     viii) The analyzer operator  shall  be  able  to select  one of the
     three lower limits.

     ix)  The  analyzer  shall be  prevented  from  reading auto exhaust
     until one of the three limits is selected.

     x)   If   a  printer  is  used,   the  CO   limit  for the  test  shall be
     printed.

4.  Printer  (Optional  features  for decentralized systems, Print system
optional for centralized systems)

     i)  The  printer  shall  print the  following  for each  vehicle tested

          1) Date

          2) Vehicle license plate number

          3) Vehicle model year

          4) Applicable cutpoints or standards  for HC and CO (standard
          feature for decentralized systems)

          5) Integrated vehicle test values  for HC and CO (standard
          feature for decentralized systems)

          6) A pass or fail indication (standard feature  for decentra-
          lized systems)

     ii) Items  1)  to 3) in 4. i) may be deleted if each  vehicle tested
     is assigned a sequential number by the  printing system that can be
     transferred to the inspector's invoice  or  testing form.

     iii) Appropriate means  (switches etc.) shall be provided  to enter
     the data  required in 4.i) 1) through 4).

     iv) The system shall determine items 4.i)  5) and 6).

-------
                               40
5.  Automatic Data Collection (optional for all systems)

     i) The  automatic data  collection feature shall  store  all of the
     pertinent data on a magnetic cassette tape.

     ii) The tape shall contain data on at least 150 vehicles.

     iii)  The tape shall be available only to an appropriate inspector,
     and shall be protected by a seal.

     iv) A keyboard  shall  be available to allow the following  types of
     data to  be entered.   As indicated,  the  system may automatically
     enter certain data:

          1) Date (Auto)

          2) Vehicle license plate number

          3) VIN

          4) Vehicle make

          5) Vehicle model

          6) Vehicle Model Year

          7) Emission Family Number (from emission label)

          8) Odometer

          9) Test Facility I.D. (Auto)

          10) Facility Test Number (Auto)

          11) Applicable cutpoints for HC and CO (cutpoints may be
          automatically determined and entered based on Vehicle Model
          Year selection)

          12) Anti-dilution decision criteria  (ie. air pump, no air
          pump, etc.)

     v) The processor shall then enter the following data on the tape.

          1) Integrated vehicle test values for HC and CO.

          2) A pass or fail indication for HC and CO.

-------
                           41
System Diagnostic Testing

 a) Switches  or  other devices for rendering any fail-safe or auto-
 matic feature inoperative for the purpose of diagnostic or perfor-
 mance checking of the analyzer are permitted.

 b) These switches (or devices) must be contained in a tamper proof
 box(es)  or  other  tamper-proof  mechanism with  provisions  for a
 seal.

 c) All  analyzer systems  must be  shipped with all  fail-safe  and
 automatic  features  operating,  and  the defeat  systems  sealed.

-------
                                    42
I.  System Correlation (on Raw Exhaust) to Laboratory Analyzers

     1.  All analyzers (comparison to NDIR Analyzer per Test Procedure)

          a) Precision Test:       A P 1 5%
          b) Slope Test:            0.95 < m < 1.05
          c) Ratio of Nodal Ave:   0.90 7R 7 1.10

     2.  HC analyzer (comparison to FID per Test Procedure)

          a) Analyzers produced through December 1985:

               i) During  analyzer check  out  determine  the  following para-
               meters between  the  candidate  instrument and an FID (See Test
               Procedure)

                    1) Precision
                    2) Slope comparison
                    3) Patio of Modal Averages

               ii) Each analyzer  manufacturer  shall use the data to develop
               a historical  comparison  to FID analyzers that can be used to
               evaluate,  prior to implementation, the correlation parameters
               specified  for  analyzers  produced  after  January  1,  1986.

          b) Analyzers produced  after  January 1, 1986 (see Test Procedure):

               i) Precision Test:           A P   1 5%
               ii) Slope Test:        0.317 1 m ± 0.350
               iv) Ratio of Modal Ave:  1.80 1 R 1 2.20

-------
                                    43
J.  Manuals
     1.  Each  analyzer shall  be delivered  with one  each  of the  following
     manuals.

          a) Operating Instructions

          b) Maintenance Instructions

          c) Initial Start-up Instructions

     2.  The  manuals shall  be constructed  of  durable materials,  and shall
     not deteriorate as a  result of  normal use  over a  five  year period.

     3.  The operating instructions must contain   :

          a) The analyzer model and serial number

          b) The propane/hexane conversion factor

          c) A step by step sequence of pre-test procedures  (span,  leak
          check, probe insertion, etc.)

          d) Sampling procedures

          e) A step by step sequence of post-test  procedures  (hang-up etc.)

     4. The maintenance manual must contain:

          a) Name,  address,  and phone number of the manufacturer's customer
          service  and  maintenance  center  at   the  home  office  and nearest
          field office.

          b) Name,  address,  and phone number of  the  nearest service center
          authorized to make warranty adjustments.

          c) A technical description of the  system.

          d) A  separate  section that clearly outlines  the required or anti-
          cipated  maintenance  schedule.   The schedule shall be broken down
          into maintenance intervals such as weekly, monthly, etc.

          e) A  separate section  that provides a  step by step sequence for
          each maintenance requirement.

          f) A  list of  replaceable items such  as  filters, probes,  etc. with
          part  numbers,  and the  estimated   service  life  of each component.

          g) A  list of  recommended  spare parts  that  the user should main-
          tain.

          h) Functional mechanical and electrical  schematics.

          i) The manufactuer's warranty provisions.

-------
                               44
i. Each manual  shall  he  attached  to  the  analyzer in a manner that will:




    a) allow convenient storage,




    b) allow easy  use,  and




    c) prevent accidental  loss  or destruction.

-------
                                    45
V.  Test Procedures

The  test  procedures in  this chapter are  presented  to provide a consistent
interpretation of  the  specifications.   Although the procedures do provide a
step by  step sequence,  and it  is  expected  that  the  evaluator follow  that
sequence,  they  are not  a substitute  for  common  sense  or good engineering
judgment.  Some  analysis system designs may  simply  not  be amenable  to  cer-
tain portions  of certain  test procedures.   Further,  there  may  be ways to
simplify or  combine certain test procedures.  In  those cases,  the evaluator
should review  the  procedure as written, determine the important conceptual
aspects and  parameters  of the procedure,  and  then use engineering  judgment
in testing the analyzer.

The basic test procedure should be performed in a  laboratory  that has  exper-
ience  in  automotive emission  testing.   The general  testing  should be  con-
ducted at  an ambient  temperature between  68°F and 86°F.   Section G  requires
certain  test  procedures  to be  conducted at  other  ambient  temperatures.

If either  of the  non-gas spanning  alternatives  are  used (See Chapter  IV,
Section  D.7.),   only  one  gas  span  operation is  permitted  for  the  entire
Chapter  V  procedure,   unless  otherwise  noted.    This single  gas   spanning
operation  shall  take  place during  the initial start-up  procedure.  Elec-
trical  span  will  be  substituted  for  all other  gas  spanning operations.
Analytical gases,  of  course, will be required  for check  out  of the  system.

-------
                                    46
A.  Traceability of Analytical Gases

The traceability of  analytical gases may or may not be part of  the  analyzer
manufacturer's  responsibility.   The choice belongs  to the state.   However,
we would  recommend  that the analyzer manufacturers not be held  responsible.
We suggest  that gas manufacturers meeting  certain performance standards be
accredited  as  acceptable  vendors,  or  that the state provide  distribution
service.

The traceability  of all gases  used  in the I/M program  is of course  impor-
tant.   Traceability  implies  not only blending, but analyzing  the gases pro-
perly.   Although  operating  the  analysis   equipment  properly  seems like  a
trivial  task,  some very  prominent  laboratories have  sometimes been  embar-
rassed  by  subsequent   review   of  their  procedures and  analysis  of  their
blended gases.

In order  to  provide the automobile manufacturers  with what EPA/OMSAPC con-
siders  proper  traceability  procedures,  the  Emission   Control Technology
Division  (ECTD) of EPA is publishing a Recommended Practice for  Assuring Gas
Traceability to NBS.  This procedure, which is a compilation of  internal EPA
procedures,  is  applicable to I/M analytical gases.

The entire  traceability procedure  may involve more  equipment than a  state
may wish  to become  initially   involved with.   A  minimum recommendation for
I/M programs   is  that   each  gas  blender or  certifying  laboratory  that is
allowed  to  supply  gases  for pass/fail  I/M systems should  follow  the ECTD
procedure.  The next step  would be a  continuing  quality-audit  of each sup-
plier by  the  state using the correlation portion  of the  traceability  proce-
dure.

-------
B.  Gas Cylinder Specifications




No test procedures required.

-------
C.  Durability Test Procedures

     1.  Vibration and Shock Test Procedure

     The  vibration and  shock  test procedure  is designed  to evaluate  the
     ability of the analyzer to tolerate moderate  use  in  a  service  facility.
     Although a simple span test is performed,  the real  test  of  the analyzer
     is the ability of the analyzer to meet  the remaining performance  checks
     to follow this test.

          a) Equipment Required

               i) Candidate instrument.

               ii) One  span  gas between 70  and 90 percent of  full scale on
               the low range, and bottled  zero  gas.

               iii) A  special  test floor  6  feet  by  10 feet.   The top  of the
               floor shall be elevated 2 inches off  the  test  facility  floor.
               The  floor  shall  consist of  an expanded  metal grating  with
               diamond  shape  openings measuring   1  x  3.7 inches or equiva-
               lent.  The  length of the  floor  in  the direction of the  "short
               way of the  diamond"  shall be  10  feet.

          b) Test Sequence

               i) Warm up  the analyzer.

               ii) Span  the  analyzer on the low  range with the  span gas per
               the  manufacturer's  recommendations   (on   the  test floor).

               iii) Record the zero response and  the span response.

               iv) Roll  the  instrument the  entire length of  the  test  floor
               in  the  direction of the  "short  way of  the  diamond", and off
               the end of  the test  floor onto  the  facility  floor.

               v) Pull the analyzer back onto  the  test floor.

               vi) Repeat  steps iv) and  v) a total of  six times.

               vii)  Check  the  zero response  and  span response   (do  not
               adjust the  analyzer).

               viii) Record the zero response  and  the  span  response.

          c) Calculations

               i)  Subtract the zero response  from the span response in step
               b) iii)	(span before).

               ii) Subtract the zero response  from the span response in step
               b) viii)	(span after).

-------
                          49
     iii) Subtract the zero response  in  step  b  viii)  from the zero
     response in b) iii)	(zero shift).

     iv) Subtract  the value  calculated  in  step  c)  ii)  from  the
     value calculated in step c) i)	(span shift).

d) Acceptance Criteria

     i) If the values calculated in step  c  iii)  and  c iv)  are less
     than or  equal  to the vibration  and  shock  zero  and  span shift
     specifications,  then  the  vibration and  shock  performance  is
     acceptable.

-------
                               50
2. Sample Line Crush Test Procedure

The  sample  line  crush  test is  to  be  repeated  under the  "operating
environment"  testing.    If  space is  a  problem  in  the  environmental
chamber, after  stabilization,  the sample  line  can be  removed  from  the
chamber  to  perform the  crush test.   If  the  line is  removed  from  the
chamber, the  crush  test must be  performed and  the line returned  to  the
chamber within 3 minutes.

     a) Equipment Required

          i) Candidate instrument.

          ii)  One span gas between 70  and  90 percent  of full  scale on
          the low range, and bottled  zero  gas.

          iii) One vehicle with at least 4000 pound  curb weight.

     b) Test Sequence

          i) Stabilize the sample line at  the prevailing ambient
          temperature.

          ii)  Warm up the analyzer.   Do  not turn on  the sample  pump.

          iii) Span the analyzer  with  the  analytical  gases.

          iv)  Leak-check the system.

          v) Stretch out the sample  line across a solid (concrete etc.)
          floor.

          vi)  Drive across the sample  line so that at least  one front
          and one rear vehicle wheel  passes over the  sample  line.

          vii) Back over the sample  line so that at  least  one front  and
          one rear vehicle wheel  passes  over the sample line.

          viii) Repeat steps vi)  and  vii)  twice.

          ix)  Leak-check the system.

          x) Check for "low  flow" indication.

          xi)  Check for visible failures,  kinks,  .deterioration, etc.

     c) Calculations
          (none)

     d) Acceptance Criteria

          i)  If  the system  passes  the  leak  check (step  ix),  does  not
          indicate a  "low  flow" condition (step x),  and  shows no sign
          of damage (step xi), then  the  crushability  performance  of  the
          sample line is acceptable.

-------
                               51
3.  Sample Handling Temperature Effect

     a) Equipment Required

          i) Candidate instrument.

          ii) Catalyst equipped vehicle.

          iii) Thermocouple read-out device.

          iv) Type "J" or type "T" thermocouple  (1/16 inch diameter
          MgO shielded and grounded).

          v) One HC span gas between 70 and 90 percent of full scale
          on the low range, and bottled zero gas.

          vi) Associated fittings.

     b) Test Sequence

          i) Locate a  point  approximately 1 inch upstream from  the end
          of  the  probe,  and  weld a thermocouple  boss  on the extender
          (top side when installed on vehicle).

          ii) Install  the  thermocouple  into the extender to a distance
          approximately  half-way  between  the  extender  wall  and the
          sample probe.

          iii) Warm up the analyzer.

          iv) Span the analyzer with the analytical gases.

          v) Leak check the system.

          vi) With the  vehicle   running,  locate a position  away from
          the  vehicle  to  avoid  vehicle  contamination,  and   measure
          the  background HC  levels  with  the sample  system.    Record
          the background HC levels.

          vii) Install the extender on the vehicle.

          viii)  Adjust the  vehicle to elevate the exhaust gas tempera-
          ture  (as measured  by  the  thermocouple)  to at  least  300°C
          (572°F)and not more than 360°C  (680°F).

          ix) As  soon  as the exhaust gas  has  stabilized  between  300°C
          and 360°C, start a timer and begin sampling.

          x)  After  twenty  (20)  minutes  at  the  temperature  in step
          viii), remove the probe or extender, and simultaneously  start
          another timer.

-------
                          52
     xi)  Immediately  move  the  probe or  extender to the  location
     previously used to measure background HC, while continuing  to
     sample room air.

     xii)  Twenty  (20)   seconds  after  the  probe or  extender  is
     removed  from the  tailpipe,  record  the analyzer's response.

     xiii)  Introduce  zero  gas  into the  analyzer through  the gas
   '  spanning system.  Record the response.

     xiv)  Introduce  span  gas into  the analyzer  through   the gas
     spanning system.  Record the response.

     xv) Leak check the system.

     xvi) Check for "low flow" indication.

     xvii)  Check  for visible  failure,  melted  or deformed  parts,
     deterioration etc.

     xviii) When the testing is completed, remove  the thermocouple
     and cap the boss with a leak proof cap.

c) Calculations

     i)  Compute  the difference  in  zero  response between  step  b)
     xiii) and b) xiv) 	  (hang-up  HC  plus  background  HC  and  zero
     drift).

     ii)  Compute  the difference in  zero  response between  step  b)
     xiii) and b) vi) 	  (hang-up HC minus  background).

     iii) Compute  the  difference in span  response between step  b)
     xv) and b) iv) 	  (span drift).

d) Acceptance Criteria

     i)  The  value  computed in c) i) must  be less  than  the hang-up
     specifications plus the zero drift specifications.

     ii)  The  value computed in c) ii)  must  be  less  than  the hang-
     up  specifications.

     iii) The value computed in c) iii) must be  less than  the  span
     shift specifications.

     iv) The system must pass the leak  check.

     v)  A low flow condition shall not  be  indicated.

     vi) No portion of  the  system shall show signs of heat damage.

     vii) If the above criteria are  met,  then the high  temperature
     performance of the  system is acceptable.

-------
                               53
4.  Filter Check and Hang-up Test Procedure

The filter check procedure consists of  two phases - a  2 hour  check,  and
a  check  to determine  the useful  life of  the  filter.  It is  intended
that all  subsequent  system check procedures are to be conducted  with  a
filter  and sample  system that  have  experienced  at   least  2 hours of
exhaust  sample  as  defined in  this  procedure.  Therefore,   the  second
part  of  the  filter  test  may  be performed  after  the other  tests  are
complete  if a  log  of sampling  time  during  the  other  tests is main-
tained.  If during subsequent testing,  a  low flow indication  occurs  due
to filter loading, the filter may be changed, the filter  life recorded,
and testing may continue with a  new filter.

     a) Equipment Required

          i) Candidate instrument.

          ii) Test  vehicle that can create  a  hydrocarbon sample of at
          least 1500  ppmh  with a spark plug  wire  removed or the choke
          partially closed.

          iii) Clean sample filter.

          iv) One  span gas between 70  and  90  percent of  full  scale on
          the low range, and bottled zero gas.

     b) Test Sequence

          i) Install a clean sample filter.

          ii) Warm up  the  analyzer.

          iii) Span the analyzer with  the analytical gases, and  leak
          check the system.

          iv) With the vehicle running, locate  a position  away  from
          the vehicle  to avoid vehicle  contamination,  and  measure
          the background HC levels with the sample system.  Record  the
          background levels.

          v) Attach the tailpipe extender to the vehicle.

          vi) Insert the probe into the extender and begin sampling.

          vii) Begin the vehicle malfunction.

          viii) As soon as the emissions  exceed 1500 ppmh, begin  a  timer.

          ix) Maintain the HC level above 1500  ppmh, and  sample  the
          vehicle for  two  (2) hours.

-------
                          54
     x) During the two hours monitor the "low flow" indication.

     xi) After two hours, remove the probe or extender, and
     simultaneously start another timer.

     xii) Immediately move the probe or extender to the location
     previously used to measure background HC, while continuing to
     sample room air.

     xiii) Twenty (20) seconds after the probe or extender is
     removed from the tailpipe, record the analyzer's response.

     xiv) Introduce zero gas into the analyzer through the gas
     spanning system.  Record the response.

     xv) Introduce span gas into the analyzer through the gas
     spanning system.  Record the response.

     xvi) Leak check the system.

     xvii) Alternatives:  1) at this point the filter testing
     may be suspended in order to complete the other tests.  A
     log book of sampling time must be kept  for the other tests,
     or 2) reinsert the probe into the extender, and continue
     testing until the "low flow" system activates, or 3) repeat
     the 2 hour filter check at the end of the performance test-
     ing of the analyzer, then continue testing until the "low
     flow" system activates.

     xviii) Record the elapsed time to "low  flow" indication.

c) Calculations

     i) Compute the difference in zero response between step b)
     xiii) and b) xiv) 	 (hang-up HC plus  background HC and zero
     drift).

     ii) Compute the difference in zero response between step b)
     xiii) and b) vi) 	 (hang-up HC minus  background).

     iii) Compute the difference in span response between step b)
     xv) and b) iv) 	 (span drift).

d) Acceptance Criteria

     i) The value computed in c) i) must be  less than the hang-up
     specifications plus the zero drift specifications.

     ii) The value computed in c) ii) must be less than the hang-
     up specifications.

     iii) The value computed in c) iii) must be less than the span
     shift specifications.

-------
                     55
iv) A low flow condition shall not occur during  the  2 hour
check.

v) If the above criteria are met, then  the performance  of
the filter is acceptable.

vi) If the total elapsed time to "low flow" activation  is
not less than 70 percent of the analyzer manufacturer's
estimate of sample life-time of the filter, the  manufac-
turer's estimate is acceptable.  If the elapsed  time is less
than 70 percent of the estimate, the manufacturer's  estimate
shall be revised downward.

-------
                                    56
D.   Design Requirement Inspection and Test Procedures

     1.   Useful Life;  No test procedure.

     2.   Name Plate;  Visual observation.

     3.   Sample System;

          a) Visual observation.

          b) Visual observation.

          c) Visual observation.

          d) See "Analyzer Gaseous Interference and Noise Test Procedure,"
          Chapter V,  Section E.8.

          e)   i) Statement by analyzer manufacturer on partical size and
               element lifetime.

               ii) See "Filter Check and Hang-Up Test Procedure."

     4.   Sample Pump:  No test procedure.

     5.   Sample Probe;  Test procedures self-explanatory or features can
          be determined by visual observation.

     6.   Sample Line;

          a)  Flexibility  Test  Procedure:   Perform  the normal  motions re-
          quired to  use the  sample line for testing vehicles.  Perform this
          test  for all environmental  conditions  used  for check  out under
          Chapter  V   procedures.   Make  a determination  about  whether  the
          sample line  can  be used without a great deal of difficulty at the
          conditions   tested.   If it  is  determined that the  sample  line is
          not sufficiently flexible,  then the manufacturer may suggest more
          objective test procedures  and/or more data to demonstrate compli-
          ance.

          b) Self-explanatory Test Procedure.

          c) Visual observation.

          d) Statement by analyzer manufacturer.

-------
                               57
7.    Analyzer Spanning System Test Procedure

     a)  The  performance  and  acceptance  of  the  spanning  system will
     generally be  determined by the system's  performance  on the other
     test  procedures  conducted  during  the evaluation  procedures,  and
     visual  observation  of  the  equipment  features.   In  other cases
     (such as event  timers), the test procedures  are self-explanatory,
     and  are  not  listed.   Guidance  for  the  one exception,  gas span
     versus - electrical  span gain  adjustments  is  provided  by the fol-
     lowing procedure.

     b)  Equipment Required

          i)  Candidate instrument.

          ii)  One span gas 70 to 90 percent of full scale on  the  low
          range,  and zero gas.

     c)  Test  Sequence

          i)  Warm up the analyzer.

          ii)  Span the analyzer with the analytical gases.  Record
          the span value.

          iii) Perform an electrical span check.   Do not adjust
          the analyzer.  Record the electrical span point.

          iv)  Reintroduce the span gas.  Do not  adjust  the  analyzer.
          Note the span response.

          v)  While the span  gas is flowing through the  analyzer,
          arbitrarily increase the analyzer response (gain  setting)
          to a value approximately 10% to 15%  greater than  the span
          value.

          vi)  Perform an electrical span check.  Do not adjust the
          analyzer.  Record  the electrical span  point.

          vii) Reintroduce the span gas.  Arbitrarily reduce  the
          analyzer response  (gain setting) to  a  value approximately
          10% to 15% less than the span value.

          viii) Perform an electrical span check.  Do not adjust  the
          analyzer.  Record  the electrical span  point.

          ix) Repeat steps ii) through viii) a total of three times.
          The increased and  decreased span adjustments  are  to be
          random settings.

-------
                          58
d) Calculations

     i) Compute the standard deviation(s) and mean(x) of  the
     electrical responses recorded in steps c) iii) , vi), and
     viii).

     ii) Multiply the standard deviation(s) by a K  factor of 2.6.

     iii) Determine the following:

          y  = x + Ks
          y2 = x - Ks

     iv) Compute the uncertainty  (U) of the electrical spanning
     system by:
          TT =            sPan Point - y     .ftn
              Electrical Span Point
     where, 1=1,2

e) Acceptance Criteria

     i) If the uncertainty is less than or equal  to  the calibration
     curve uncertainty (Chapter IV, Section E) , the  electrical
     span system is acceptable.

-------
                               59
8.   Analyzer Ranges:  Visual observation.

9.   System Grounding;  See Test Procedure E.9. (Analyzer Electrical
     Interferences).

10.  System Vents; If  restrictions  downstream of the analyzer exit are
apparent, a test procedure shall be devised that evaluates the analyzer
performance-under restricted conditions.  Test Procedures E.3.a) (Alti-
tude Compensation) and F.I.  (Sample Cell Pressure Variation, Low Flow,
and Response Time) shall be used for guidance.

-------
                                    60
E.  Analyzer Performance Test Procedures

     1.  Calibration Curve Test Procedure

          a) Equipment Required

               i) Candidate instrument

               ii) Seven or more calibration gases for each range of each
               analyzer roughly equally spaced over each range.

               iii) Zero gas

               iv) Associated valves and fittings

          b) Test Sequence

               i) If necessary, follow the manufacturer's instructions  for
               initial start-up and basic operating adjustments.

               ii) Warm up the analyzer

               iii) Zero the analyzer with the zero gas

               iv) Span the analyzer with one of the calibration gases.  The
               span point  should  be approximately 80  to  90 percent of  full
               scale of the low range

               v) Recheck  the  zero.  If the zero  has  shifted, repeat  steps
               iii), iv),  and v) a maximum of one more time.

               vi) Do  not  adjust  the zero or  span controls on the analyzer
               for the remainder of the test.

               vii) Introduce  the  calibration  gases  in  ascending  order of
               concentrations  beginning  with  the  zero  gas.  Record the
               analyzer response to each concentration value.

               viii) After the highest concentration has been  introduced and
               recorded, introduce  the  same  calibration gases to the analy-
               zer in a descending order.  Include the zero gas.  Record the
               response of the analyzer to  each  gas.   Record negative  zero
               responses (if any) as they occur as negative values.

               ix)  Repeat  steps  vii),   and  viii)  a  total  of  five  times.

          c) Calculations

               i)  For  hydrocarbon  analyzers,  compute  the hexane equivalent
               (ppmh)   of   each  calibration  gas  by multiplying the concen-
               tration  value  in  ppm propane  (ppmp)  by  the  propane/hexane
               conversion  factor listed on the analyzer.

-------
                          61
     ii)  Compute  the  standard  deviation(s)  and  mean(x)  of  the
     analyzers  response  for  each  concentration   of  calibration
     gases.   Include  both  upscale   responses and  downscale  re-
     sponses  to  the  same calibration gas.   Zero may be a  special
     case and not amenable to the calculations.

     iii)  For all concentration values  except the highest value,
     multiply  the  standard  deviation(s)  by a factor  (K)  of 2.5.

     iv)  Multiply  the  standard  deviation(s)  of  the  analyzers
     response  to  the  highest concentration  by a K factor  of 3.5.

     v) Compute the following for each concentration

          1) y  = x + Ks

          2) y2 = x - Ks

     vi)  Compute  the uncertainty(U)  of  the calibration curve  for
     each concentration by:

              concentration value-y.      n
          U —            .     ,    i  x  lUu
              concentration value

    where i = 1, 2

d) Acceptance Criteria

     i) Identify the maximum uncertainty  for each  range.

     ii)  If  the  maximum uncertainty  is  less than  or equal to  the
     uncertainty  specification  (Chapter  IV, Section E),  (plus  or
     minus),  the  calibration curve  is  acceptable.  If the uncer-
     tainty  is  greater  than  specification,  the calibration  curve
     is not acceptable.

     iii)  If  the  calibration curve  is  not acceptable,  then  the
     instrument manufacturer should undertake  an engineering  study
     to  identify  the  cause of  the problem  prior  to  continued
     testing  or  introduction  of  the analyzer  to the commercial
     market.

     iv)  After  the  cause  of  the problem   is  identified  and  the
     analyzer  is repaired  or adjusted   this  test  should  be  re-
     peated.

e) Repeat steps b),  c), and  d)  for each  range of the  analyzer.

-------
                               62
2.   Analyzer Resolution

     a)  Analog Meters:  The  resolution is determined by the interval of
     the smallest graduation of the meter face for each analyzer range.

     b)  Digital  Meters:  The resolution  is  determined  by  the increment
     of   the  least  significant digit  of  the meter  readout  for  each
     range.

-------
                               63
3.   Compensation Test Procedures

     a) Altitude Compensation Test Procedure   (not required for analyzers
     with Spanning Alternative, 1 or 2)

          i) Eauipment

               1) Candidate instrument.

               2) One pressure gauge or manometer that reads  in absolute
               pressure (e.g. 20 to 32 in. HgA) with  1% accuracy.

               3) One differential pressure gauge (0  to 15 inches of
               water).

               4) One flowmeter (0-20 cubic feet per  hour  (CFH)).

               5) Two needle valves.

               6) One vacuum pump (greater than 20 CFH'? 24 in. HgA).

               7) One span gas 70 to 90 percent of full scale on the  low
               range.

               8) One Tedlar sample bag.

               9) Associated lines and fittings (non-reactive).

          ii) Test Sequence

          This procedure  is written  around  parallel  HC  and CO analy-
          zers,  and  should be  performed  on  each  analyzer.   If  the
          analyzers  are  in a  series  configuration,  then an  additional
          pressure  gauge(s) will   be  required,  but  the  test  can  be
          performed on both analyzers at  the same time.

          If  the  analyzer manufacturer  states  in  writing  that  the
          structural  integrity of  the  sample  cell  will  not withstand
          the pressure differential associated with  this test procedure
          at  24  inches of  HgA,  the altitude  check  must  be  made in  an
          altitude chamber.  In  an altitude chamber  only steps 8), 9),
          10), 14), 15), 16) and 20) need be performed at  the pressures
          specified in step 13) and 18).

               1) Identify  the sample line entering  the sample cell  of
               each analyzer.

               2) Install  a tee fitting  in the sample line immediately
               upstream  of the  sample cell   (as close  as practical).
               Install the  tee with branch pointing  up.

               3)  Immediately  upstream of  the tee  install  one of the
               needle valves.

-------
                64
4)  Identify the  exhaust  line  from  the sample  cell  and
install  two tees  or  a 4-way  cross  fitting in  the  line
with the branches  pointing  up.

5)  Downstream  of  the  tees  or  cross  fitting install  the
flow meter.

6)  Attach  a  differential   pressure  gauge to  the  tee
upstream of the  sample cell  and  one  of the  exhaust  tees.
Attach  one gauge  across each  analyzer in  a series  sys-
tem.

7)  Attach  the  absolute  pressure gauge to  the  second
exhaust tee.

8) Warm up  the analyzer.

9) Check for leaks.

10)  Sample  room air  through  the sample  line  with  the
needle valve wide  open.

11)  Record the  sample cell  exhaust  flow rate,  the  dif-
ferential  pressure across  each  analyzer,   and the  abso-
lute pressure.

12) Attach the second needle  valve  after  the  flowtneter.

13) Adjust both  the needle  valves to  obtain an  absolute
pressure  of 31  inches of  HgA,  the  sample differential
pressure  recorded in  step  ii),  and  a flow-rate of  not
less than  2 cfh.

14) Fill the sample bag with the span  gas,  and  introduce
the span gas from  the  bag into  the sample  line.

15) Use the analytical gases to  set  the zero and span of
the  analyzer.    Note   if  internal adjustments   were  re-
quired to  span the analyzer.

16) Return  to sampling room  air.

17) Attach  the vacuum  pump  to  the exhaust  line  after the
second needle  valve  with an appropriate  length  of  line,
and open both needle valves.

18)  Turn  the  vacuum  pump   on,  and   adjust both needle
valves  to  obtain  an  absolute  pressure of 24 inches of
HgA,  the  same   differential  pressure  recorded  in  step
11), and  a flow rate  approximately  the same as  recorded
in step 11) .

-------
                     65
     19) Refill  the sample  bag  if  necessary,  and  introduce
     the span gas into the sample  line  from  the bag.

     20) Use the analytical gases  to set  the  zero  and  span  of
     the analyzer.   Note  if  internal  adjustments  were re-
     quired to span the analyzer.
iii)  Calculations  (none)

iv) Acceptance Criteria

     1)  If the analyzer can be spanned properly  in  steps  ii)
     15), and  ii)  20,  and the technique required  to  span  the
     analyzer  under  these  conditions  is   identified in  the
     owner's manual,  then the  altitude compensation network
     is acceptable.

-------
                          66
b)    Pressure and Temperature Compensated Analyzer Test Procedure

     i) Testing  concepts:   This  test procedure is to be performed
     in  order  to  identify the  performance  of  any  pressure or
     temperature  compensation  systems under  the  various environ-
     mental  conditions  that  may   be  encountered  during  vehicle
     inspection  testing.   In  general,   temperature  compensation
     will  be  evaluated during the  basic  check-out  (68°F to 86°F)
     and during  the more  hostile  environmental temperature tests
     specified  in  Section  G.   No other  special  testing  would
     normally be necessary.

     In  order to  evaluate  pressure compensation  systems,  addi-
     tional  testing is  necessary.   If the  analyzer manufacturer
     can make a case that  testing the pressure  compensation  system
     in a  manner  similar to the procedure  specified  in Chapter V
     Section E.S.a)  (altitude  compensation) will represent  actual
     analysis system operating  conditions in  the field, then  that
     procedure (E.3.a) may be used  for check-out.  If a sufficient
     case  cannot  be made,  and a  suitable alternative test  proce-
     dure  cannot be determined,  then performance  evaluations of
     the pressure  compensation system  must be  carried  out  in an
     altitude  chamber.  The pressure  compensation  test  shall be
     conducted at  each environmental temperature condition  speci-
     fied  in Chapter  V.   The  tests  shall be performed  on  each
     range of each analyzer.

     ii)   Test  conditions:  The  test  conditions  shall consist of
     three  basic barometric pressures  about  which  excursions in
     pressure shall be made.  The values  are:

          Basic Test                Excursion
          Point  (inches HgA)        Points (inches HgA)

               24.5                  24.0, 26.5
               28.5                  26.5, 29.5
               30.0                  28.5, 31

     iii)  Test Sequence

           1)  Identify the  concentration  value  that  provides  the
          greatest  uncertainty  in  measurement as  determined by
          Test Procedure E.I).  Obtain a calibration gas at  that
          level  (same  bottle if possible).

          2)  Set  up the test equipment for one of the three  test
          points.

          3) Warm up the analyzer.

          4)  Gas span  the analyzer with a different calibration
          gas between  80 and 90 percent  of full  scale on the low
          range and bottled  zero gas.

-------
                67
5) Leak check the system.

6)  Alternately  introduce  zero  gas  and  the calibration
gas identified in step 1) through the sample prohe until
a  total of  three  readings are obtained and recorded  for
each test step.

7)  The  test  steps  for the 24.5  in  HgA  test point are:

     Test Step           Action

     24.5                gas span
     24.0                3 readings
     24.5                3 readings
     26.5                3 readings
     24.0                3 readings
     26.5                3 readings
     24.5                3 readings
     24.0                3 readings
     24.5                3 readings
     26.5                3 readings

8) Repeat steps 4) through 6) at  the 28.5  inch HgA test
point.  The test steps for this test point  are:

     Test Step           Action

     28.5                gas span
     26.5                3 readings
     28.5                3 readings
     29.5                3 readings
     26.5                3 readings
     29.5                3 readings
     28.5                3 readings
     26.5                3 readings
     28.5                3 readings
     29.5                3 readings

9)  Repeat steps 4) through 6) at  the 30.0  in. HgA  test
point.  The test steps for this test point  are:

     Test Step           Action

     30.0                gas span
     28.5                3 readings
     30.0                3 readings
     31.0                3 readings
     28.5                3 readings
     31.0                3 readings
     30.0                3 readings
     28.5                3 readings
     30.0                3 readings
     31.0                3 readings

-------
                     68
iv) Calculations

     1)  For  each  test  point,  compute  the  standard  devia-
     tion (s) and mean  (x) for  the 9 readings  obtained  at  each
     pressure level.

     2)  Multiply each  standard deviation  by a  K factor  of
     2.6.

     3)  For  each pressure level at  each test point,  compute
     the following:

          y. = x + Ks

          y2 = x - Ks


     4)  For  each pressure level at  each test point,  compute
     the  uncertainty  (U)  of  the  measurements  based  on  the
     calibration gas used in step iii) 6) from:

              concentration value-v.     ,nr,
          U = 	:	:	"-f  x  100
              concentration value

     where i = 1, 2

v) Acceptance Criteria

     1) Identify the maximum uncertainty

     2)  If  the  maximum uncertainty is  less  than or equal  to
     the uncertainty specification  (Chapter  IV, Section  E),
     (plus  or  minus),  for   the basic  analyzer  calibration
     curve, the  pressure compensation is acceptable.   If  the
     uncertainty is  greater  than specification,  the pressure
     compensation is not acceptable.

     3)  If  the  pressure compensation is not  acceptable,  then
     the instrument  manufacturer should  undertake an  engin-
     eering study  to identify the cause of the  problem prior
     to  continued  testing  or  introduction of the  analyzer to
     the commercial market.

     4)  After  the  cause of the  problem  is identified  and  the
     analyzer  is repaired  or  adjusted  this  test should  be
     repeated.

vi) Repeat steps iii),  iv), and  v) for each range  of  the
analyzer.

-------
                          69
c) Test Procedure for Analyzers with Improved Accuracy

     i) If  an analyzer  manufacturer claims  that  the measurement
     uncertainty  specifications (Chapter  IV,  Section H.)  can he
     met  without  pressure  and/or temperature  compensation,   this
     claim  can  be verified  by the  test  procedure  in  Section K.
     3-6)  of this Chapter.

-------
                               70
4.   Analyzer Zero and Span Drift Test Procedure
    (See Test Procedure Section F.I.a), Warm-up Lock-out Test Procedure)

     a) Equipment Required

          i) Candidate instrument.

          •ii) One  span gas  between 70 to 90 percent  of  full scale on
          the low range.

          iii) One cylinder of zero gas.

          iv) One chart recorder  (one megohm impedance or greater) with
          approximately a 10 to 12  inch wide chart.

     b) Test Sequence

          i) Begin  the test  sequence with an  analyzer  that is  turned
          off, and has stabilized at the prevailing ambient  temperature
          for at least 3 hours.

          ii) Remove the analyzer's protective cover.

          iii)  Locate  the   meter  readout  and  attach  leads  from the
          readout terminals to the  chart recorder.

          iv) Reinstall the protective cover.

          v) Turn on the analyzer.

          vi) Select a voltage range on the chart recorder so that full
          scale  of  chart  recorder  equals  full  scale voltage  of the
          analyzer meter on the range under test.

          vii)  The   chart   recorder must  indicate  both  negative and
          positive zero drift.  If  necessary, offset the chart recorder
          zero 5 units up scale.

          viii)  As  soon as  the warm-up  lock-out  feature deactivates,
          zero  and  span  the  analyzer  with  the  analytical gases per
          manufacturer's  operating  instructions,  and start the  chart
          recorder at  a  minimum chart speed of 0.5  inches  per minute.

          ix) Do not  adjust the analyzer or recorder controls (zero or
          span) for the remainder of the test.

          x) Reintroduce  zero  gas  (if not done already)  and start the
          test.

          xi) Mark  the chart  paper indicating  the  zero response, the
          span response, and the start of the test.

-------
                          71
     xii) Flow zero gas through  the analyzer  for  one  hour.

     xiii) Introduce  the  span gas to  the  analyzer (do not  adjust
     the equipment).

     xiv) Reintroduce  the zero gas to  the  analyzer  (do not  adjust
     the equipment).

     xv) Reintroduce  the  span gas to  the  analyzer (do not  adjust
     the analyzer).

     xvi) Flow span gas through  the analyzer  for  one  hour.

     xvii) Reintroduce zero gas  to the  analyzer.

c) Calculations

     i) Compute  the  difference  (as a  percent of full scale  chart
     deflection) between  the  analyzer  zero response  for  the  first
     span check  (step  b)viii)) and the zero  response after  the  1
     hour span check (step b)  xiv)) 	 (zero drift).

     ii) Locate on the chart  the maximum and  minimum  analyzer zero
     response  during  the  first  1  hour  period   (step  b)  xii)).

     iii) Compute  the difference  (as  a  percentage of full  scale
     chart deflection) between the analyzer  zero-response  for the
     first span  check (step  b)  viii)), and  the maximum zero re-
     sponse  and  then  the minimum zero-response  as  identified  in
     step c)  ii) 	 (zero drift).

     iv)  Compute  the  analyzer  span-response at  the  1  hour  span
     check as the difference  between the chart  reading of  the span
     gas  (step  b)  xiii)   and the chart  reading of  the  zero  gas
     (step b) xiv)) 	 (span  response).

     v) Locate  on  the chart  the maximum and  minimum  chart  reading
     during the one hour  span  period  (step  b) xvi)).

     vi)  Compute  the  difference  (as  a percentage  of  full  scale
     chart deflection) between the analyzer  zero-response  immedi-
     ately prior  to  the   one  hour  span period  (step b) xiv)), and
     the  maximum  and  minimum  span  chart  readings  identified  in
     step c)  v) 	 (span response).

     vii) Compute  the difference  (as  a  percentage of full  scale
     chart deflection)  between  the  span  chart deflection at the
     end  of  the  1  hour   span,  and the  zero-response after  the  1
     hour span 	 (span  response).

-------
                          72
     viii) Compute  the difference  (as  a  percentage of full  scale
     chart deflection)  between  the  span  chart deflection  immedi-
     ately prior  to  the start of  the  one hour zero test  (step b)
     viii))  and  the   zero-response  just  prior  to  that   span
     check 	 (span response).

     ix) Compute  the  difference (as a  percent of full  scale  chart
     deflection) between the zero-response just prior  to the  start
     -of the 1 hour span (step b) xiv)), and  the zero-response just
     after the 1  hour span   (step  b) xvii))  	  (zero   drift).

     x) Compute  difference (as  a percentage  of  full  scale  chart
     deflection) between  the  following span responses  in  step c) :
          (vi max) to (iv)    span drift
          (vi min) to (iv)    span drift
          (vii)  to (iv)      span drift
          (viii) to (iv)      span drift
d)  Acceptance Criteria

     i) If  each value  computed in  steps  i),  iii)  max, iii) min,
     and  ix)  is less  than or  equal to the specifications  listed
     for zero drift, the zero drift  of  the  analyzer  is  acceptable.

     ii) If each value computed in  step  x) is less  than  or equal
     to the specifications for span drift,  the span drift  of  the
     analyzer is acceptable.

-------
                               73
5.   Analyzer Span Drift Test Procedure




(See Analyzer Zero and Span Drift Test  Procedure,  E.4.)




6.  Analyzer Noise Test Procedure




(See Analyzer Gaseous Interference and  Noise  Test  Procedure,  E.8.)

-------
                               74
7.  Analyzer Sample Cell Temperature Test Procedure
   (See Test Procedure Section H.I.a), Warm-up Lock-out Test Procedure).

     a) Equipment Required

          i) Candidate instrument.

          ii) Thermocouple readout device.

          iii) Type J or type T thermocouple.

     b) Test Considerations

          i)  The  sample  cell  test  procedure  may be  run concurrently
          with the  zero/span drift test or any  other  test that begins
          with an analyzer that is not warmed up.

          ii) It  is recommended  that the  test  equipment  remain hooked
          up  to  the analyzer  for the  duration  of  the  analyzer check
          out, and  the  temperature of the sample  cell  should  be moni-
          tored from time to time.

     c) Test Sequence

          i) Locate on  the  sample cell a point,  based on engineering
          judgement, that would be the coldest point.

          ii) Attach a  thermocouple  to the sample cell at the  location
          identified in step i).

          iii) Conductive,  convective,  and   radiation  losses  must be
          considered  in  the   selection  of  the   location,  and  in  the
          manner of thermocouple attachment.

          iv) Stabilize  the  analyzer at the prevailing ambient  temper-
          ature for at least 3 hours.

          v) Record the sample cell temperature.

          vi) Turn on the analyzer.

          vii) As  soon  as the  warm-up lock out  feature is deactivated,
          record the temperature of the sample cell.

          viii) Immediately switch to sample ambient air  (or analytical
          gas  at  the  ambient  temperature)  through  the  sample line.

          ix)  Monitor  the   sample cell  temperature over  the  next  5
          minutes.

          x)  Record the  lowest  sample cell  temperature  during the 5
          minute period.

-------
                          75
d) Calculations
     (None)

e) Acceptance Criteria

     i) If the temperature recorded  in  step  c)  v)  is  approximately
     the same as the ambient temperature,  the analyzer  is  properly
     stabilized.

     ii) If  the  temperatures recorded  in  step  c)  vii)  and step  c)
     x) are equal to or greater than  the specifications for sample
     cell  temperature,  the  sample cell  temperature  is  acceptable.

     iii)  While  monitoring  the sample  cell  temperature during  the
     analyzer check-out procedure, if the  temperature is less  than
     the specifications  when  concentration  data  (span  or  sample)
     is read, the acceptance in step  ii) is  void.

-------
                               76
8.   Analyzer Gaseous Interference and Noise Test Procedure

     a) Equipment Required

          i) Candidate instrument.

          ii) One  high impedence  (1  megohm  or  greater)  digital volt-
          meter with at least 4% digit resolution.  Generally, a 2 volt
          ,(1.9999 v) and  a 20 volt (19.999  v)  scale will be adequate.
          A chart recorder  of similar impedence, resolution, and scal-
          ing may be used in place of a voltmeter.

          iii) One  calibration  gas with approximately  14% CO  and the
          balance N

          iv) One steam generator.

          v) One mixing  chamber with probe attachment, dilution valve,
          and chamber air temperature readout system.

          vi) One calibration gas with approximately 100 ppm NO .  The
          NO  value must have been checked within 48 hours of the start
          of  the  testing  sequence   with  a  chemiluminescent  analyzer
          meeting the  specifications  of  40 CFP  86  Subpart B,  D, or N.

          vii) One  span  gas between  70 and 90 percent of  full scale on
          the low range, and bottled  zero gas.

          viii) Tedlar sample bags.

          ix) Associated lines and fittings (non-reactive).

     b) Test Sequence

          i) Remove the analyzer's protective cover.

          ii)  Locate   the  meter  readout  and  attach  the  voltmeter or
          chart recorder leads to the meter terminals.

          iii) Reinstall the analyzer's protective cover.

          iv) Temporarily bypass  the  water trap.

          v) Warm up the analyzer.

          vi) Introduce  the span gas to the  analyzer through the span
          port.

          vii) Adjust  the  span  so that the  analyzer reads 100 percent
          of low scale.

-------
                     77
viii) Observe the voltmeter reading at  100 percent  full  scale
of the analyzer and record the voltage  value.   Alternatively,
adjust the chart recorder so  that  full  scale  is equivalent  to
the analyzer full scale.

ix) Zero  and  span the analyzer  with  the analytical gases  to
the proper values.

x) Fill a sample bag with span gas.

xi) Introduce  the  span gas from the  sample bag to  the analy-
zer through the probe.

xii) After  the  reading has stabilized,  observe the voltmeter
or  chart  recorder.   Record the  highest value and  the lowest
value  over  a 3  minute time  span.  The scale  can  be changed
for better resolution.

xiii) Fill another sample bag with CO .

xiv) Introduce the CO  through the sample probe.

xv) Record  the average voltmeter  or  chart recorder reading.
The scale can be changed for  better resolution.

xvi)  Fill a  sample  bag  (after  purging with  N )  with  NO .

xvii) Introduce the NO  through  the sample probe.

xviii) Record  the average voltmeter  or chart  recorder  read-
ing.  The scale can be changed-for better resolution.

xix) Start the steam generator.

xx) Attach  the probe  to  the  mixing  chamber  and  adjust the
dilution  valve  to obtain  a  40°C  (101°F)  temperature in the
dilution box.  To prevent condensation  in the  analysis system
this  test should  be  performed  with  an ambient  temperature
between  30°C  and AO°C.   This test  cannot  be  performed  cor-
rectly when  testing at the  lower  ambient temperature condi-
tions  specified   in  Section  G  of  this  chapter,   and should
therefore, be omitted when testing under Section G.

xxi) Sample from the dilution box.

xxii) Record the average voltmeter or chart recorder reading.
The scale can be changed for  better resolution.

xxiii) Reconnect the water trap.

-------
                          78
     xxiv)  Repeat  steps xix) through  xxii)  with an amhient  temp-
     erature  at the  same  value  as  in  step  xx) and  40°C in  the
     dilution  box.   When  repeating  this  test  at  the  operating
     environmental  conditions  specified  in  Section  G  of   this
     chapter, perform  this  test with  a  40°C dilution  box tempera-
     ture and  the  analysis  system at  the  prevailing ambient  temp-
     erature.

c) Calculations

     i)  Compute the  difference  between  the  high and  low reading
     recorded in step b) xii), then divide that  result by the full
     scale  voltage value recorded  in step  b)  vii).   This value,
     expressed  as  a  percentage  and divided  by 2,  is defined  as
     noise of the  analyzer.

     ii)  Divide the  value  recorded  in  step  b) xv)  by  the  full
     scale  voltage value recorded  in step  b)  vii).   The  result,
     expressed  as   a  percentage,  is  the CC>   interference of  the
     analyzer.

     iii) Divide  the  value  recorded in  step  b)xviii)  by  the  full
     scale  voltage value  recorded  in  step  b)vii).   The  result,
     expressed  as   a  percentage,  is  the NO   interference of  the
     analyzer.

     iv)  Divide the  value  recorded in  step b)xxii)  by  the  full
     scale  voltage value  recorded  in  step  b)vii).   The  result,
     expressed  as  a percentage,  is  the  water interference of  the
     analyzer.

     v)  Divide  the value  recorded in  step b)xxiv)  by  the  full
     scale  voltage value  recorded  in  step  b)vii).   The  result,
     expressed  as  a percentage,  is  the  water interference of  the
     system.

d) Acceptance Criteria

     i)  If   the percentages  calculated   in  paragraph  c) for  the
     analyzer  are   equal  to or  less   than  the  specifications  for
     noise and  gaseous  interferences (Chapter V,  Section E.),  then
     the noise  and  gaseous interferences are acceptable.

     ii)  Inspect  the  analysis  system  downstream  for  condensed
     water.   If water  is  found,  the  effectiveness  of  the  water
     trap is not acceptable.

-------
                               79
9.  Analyzer Electrical Interference Test Procedure

     a) Equipment Required

          i) Candidate instrument.

          ii)  One high  impedence  (Imegohm  or  greater)  digital volt
          meter with at least 4i$ digit resolution.  Generally, a  2 volt
          (1.9999 v)  and  a 20 volt  (19.999  v)  scale will be adequate.
          'A chart recorder  of similar impedence, resolution, and scal-
          ing may be used in place of a voltmeter.

          iii) One vehicle  with  high energy  ignition system, and solid
          core ignition and coil wires.

          iv) One 3  amp  or more variable  speed  (commutator type) hand
          drill with a plastic handle.

          v) One 20 foot extension cord with  3 sixteen gauge wires, and
          a 2 plug  outlet.   For systems without ground  fault circuits,
          a  2  wire  extention  cord or  a  non-grounded adaptor  will be
          required.

          vi) A  CB  transmitter at or near FCC legal maximum power with
          a matching antenna.

          vii)  One  variable  voltage  transformer   (90v  to  130vAC).

          viii) One span gas between 70 and 90 percent of  full scale on
          the low range.

          ix) One dry flannel cloth.

     b) Test Sequence

          i) Remove the analyzer's protective cover.

          ii)  Locate  the  meter  readout,  and attach  the  voltmeter or
          chart recorder leads to  the meter terminals.

          iii) Peinstall the analyzer's protective cover.

          iv) Plug  the  analyzer's power cord into  a 2-outlet grounded
          electrical outlet box and warm up the analyzer.

          v)  Use  a  non-grounded  adapter  to check  any  ground fault
          circuit.

          vi)  If the  system  does not  have a  ground  fault  circuit,
          perform  the remainder  of  the  test  sequence  with  the non-
          grounded adapter.

-------
                      80
vii)  Mark the  adapter and  the outlet  box,  and  perform  the
testing  sequence  with  the  adapter  in that orientation.   After
completion  of  the  sequence, reverse  the  adapter orientation
(hot  lead  versus  neutral  lead)   and  repeat  the  sequence.

viii)  Introduce  the  span  gas through  the  span port,  and
adjust the span so  that  the  analyzer reads 100 percent  of low
scale.

-ix)  Observe   the  voltmeter  or  chart  recorder  reading  at  100
percent  full scale  of the  analyzer,  and record  the  voltage
value.

x)  Zero  and  span the  analyzer with  the  analytical  gases to
the proper values.   Record the  values.

xi)  Introduce  span  gas  to the analyzer.  Record  the  average
span  gas voltage  levels  over a  5 minute  time span.  The scale
can be changed  for  better  resolution.

xii)  Either  move  the analyzer or the vehicle (engine compart-
ment) to within 2 feet of  each  other.

xiii) Introduce the  span gas to the analyzer.

xiv)  Open the hood  and start the vehicle.

xv)  Record  the  average  span  gas voltage  level during  a  5
minute time  span.

xvi)  Stop the vehicle  engine.

xvii)  Plug  the electric  drill  into  the  other outlet  of  the
same  receptacle that the analyzer  is connected to.

xviii) Move   the  drill at approximately 3 to 4  feet  high to
within  12 inches of the  analyzer.  At  four  locations  around
the  analyzer  vary   the  drill speed  from minimum  to  maximum
speed.   Attempt  to locate   the  positions  that   provide  the
greatest  interference.

xix)  Record  the average span gas  voltage level at each posi-
tion.

xx) Key  the  CB  radio within  50  feet of the analyzer.

xxi)  Record   the  average  span  gas voltage  level  when  the CB
radio is  transmitting.

xxi)  Turn the analyzer off.

xxii) Plug both the  analyzer and the drill into the extension
cord.  For  systems  without  a ground fault circuit use either
a  2 wire  extention  cord,  or  a non-grounded adapter.

-------
                          81
     xxiii)  Plug  the  extension  cord  into  the  variable voltage
     transformer  and  set  the  transformer  output  to  110 volts.

     xxiv) Turn the analyzer on, and warm  it up.

     xxv)  Check the  zero and  span.   If  necessary  reset  to  the
     exact voltage level recorded in step  b) x).

   -  xxvi) Repeat step b) xii) through xxi).

     xxvii) Record the average span gas voltage  level as  indicated
     in steps b) xv), xix), and xxi).

     xxviii) Reduce  the transformer  output  voltage  to  90 volts.
     Let analyzer stabilize for 2 minutes.

     xxix)  Repeat  steps b) viii) through  xxi)  with the  extention
     cord  and  90 volt  set-up.   Non-relevant  set-up  steps may be
     omitted.

     xxx)  Increase  the transformer output to  130 volts.   Let  the
     analyzer stabilize for 2 minutes.

     xxxi)  Repeat  steps b) viii) through  xxi)  with the  extension
     cord  and  130  volt set-up.  Non-relevant  set-up  steps may be
     omitted.

     xxxii)  If  the unit  has  an analog  meter,  observe and record
     the meter reading with span gas.

     xxxiii)  Rub  the  meter  face  10  times  with  the  dry flannel
     cloth  in  the  up-scale direction.   Record  the meter reading.

c) Calculations

     i) Compute  the  difference between each set  of voltage levels
     recorded  in  step b)  x)  and b)  xv).   The maximum difference
     expressed  as a  percentage of the voltage  recorded in step b)
     ix) is defined as the RF interference (RFI).

     ii) Compute the difference between each set  of voltage levels
     recorded  in  step b)  x)  and b)  xx).   The maximum difference
     expressed  as  a  percentage of  the voltage  recorded in  step
     ,b)ix) is defined as the induction interference.

     iii)  Compute   the  difference  between  each  set  of voltage
     levels recorded in step b)x) and b)xxi).   The maximum differ-
     ence  expressed  as  a percentage  of  the voltage  recorded in
     step b)ix) is defined as the VHP interference.

     iv)  Compute  the  difference between  the  voltage  levels  re-
     corded  in  step  b) x) and step b) xxvii).   The largest result
     expressed  as a  percentage of the voltage  recorded in step b)
     ix) is defined as the line interference.

-------
                          82
     v) Compute the difference between  the voltage  levels  recorded
     in step b)  x)  and b) xxix).   Compute  the difference  between
     the voltage  levels  recorded in  step b) x) and b) xxxi).   The
     largest value of the results expressed as a percentage of  the
     voltage recorded  in  step b) ix)  is  defined  as the line vol-
     tage interference.

     vi) Compute the difference between the meter  reading  recorded
     'in step b) xxxii) and b) xxxiii).  Also compute the difference
     as a percentage  of  full scale.  The difference is defined as
     the static electricity interference.

d) Acceptance Criterion

     i) If the percentages calculated in paragraph  c) are  equal to
     or  less  than  the  electronic  interference  specifications
     (Chapter  TV,  Section E.),  then the electrical interferences
     are acceptable.

-------
                               83
LO.   Propane  to Hexane Conversion Factor Test Procedure

     a)  Testing Concepts

          i)  The conversion factor shall be evaluated at two hexane
          concentration levels.

               1)  200 ppmh ( _+ 15 ppmh) on the low scale, and

               2)  1800 ppmh ( +_ 150 ppmh) on the high scale.

          ii)  A minimum of three samples of the hexane gas at each
          concentration level shall be performed.

          iii)  Because of the exceptional hang-up characteristics of
          hexane,  an auxilliary  sample/spanning system that is specially
          constructed to be essentially hang-up free may be used.  The
          auxilliary system should be of laboratory quality, and may
          bypass the candidate analysis system controls.

          iv)  The  test should always be conducted at an ambient temper-
          ature greater than 20°C (68°F).

          v)  A flame ionization  detector (FID) properly optimized and
          calibrated . (see 40 CFR 86 Subpart D or N) may be used to
          provide  additional quality control on the conversion factor
          determination procedure.

     b)  Test  Sequence

          i)  Set up the test equipment.

          ii)  Warm-up the candidate analyzer.

          iii)  Leak check the system used.

          iv)  Span the analyzer  on the low range with a propane cali-
          bration  gas near ( _+ 5%) the expected low concentration hexane
          response, and with bottled zero gas.  Use the propane to
          hexane conversion factor indicated on the analyzer.

          v)  Alternately cycle low concentration hexane gas and the
          calibration gas through the analyzer a total of 3 or more
          times.

          vi)  Record each response for each gas.

          vii)  Span the analyzer on the high range with a propane
          calibration gas near ( +_ 5%) the expected high concentration
          response, and bottled  zero gas.  Use the propane to hexane
          conversion factor indicated on the analyzer.

-------
                          84
     viii) Alternately cycle  the high  concentration  hexane gas  and
     the calibration gas through the analyzer  a  total  of  three
     or more times.

     ix) Fecord each response  for  each gas.

c) Calculations

     i) For the values recorded in step  b) vi),  and  b)  ix),  compute
     the mean (x) and standard deviation(s)  for  the  hexane response
     and the propane response.

     ii) Use an  aof 0.05 and  the  student's  "t"  test to determine
     the confidence interval  of the population mean  at  the 90%
     confidence level based on the values  computed  in  step c)  i)
     for each range.

     iii) Use the following equation to  determine  the  mean propane
     to hexane conversion factor (CF)  for  each range:

               (.5) + (.5)(x    -  Hex  Cal)
       CF =                 hex
                            Hex  Cal

     where:

     x hex = the mean hexane  response  determined  in
             step c) i).

     Hex Cal = concentration  of  hexane calibration gas.

     iv) Determine if sufficient number of  cycles (steps  b)  v)
     and b) viii)) were run on each  range by  the  following equation:

                 (.5)(x,   -  xr_)  +  (.5) Hex  Cal
       o.oi   <_        hex    CI   ~	
                            Hex  Cal

          where:

          x hex = the mean hexane  response  determined  in  step c)  i).

          xp  = the confidence interval determined in  step c) ii).
           *-• L
          Hex Cal = concentration  of hexane calibration  gas.

     v) For each range, multiply the mean propane response (x)
     determined in step c) i) by the CF computed  in  step  c)  iii).

     vi) Determine the  difference  between the CFs for  each range,
     and determine the  mean CF.

-------
                          85
d) Acceptance Criteria

     i) If the value computed  in  step  c)  iv)  is  greater  than  0.01,
     perform additional cycles  (see steps b)  v)  and  b) viii))  until
     the computed value is  less than 0.01.

     ii) If the difference  between the CFs  computed  for  each  range
     (step c) vi)) is greater  than .030 a sperate  CF must  be  used
   ,  for each range.  If the difference is  less  than .030,  a  mean
     CF may be used for both ranges.

     iii) The CF value(s) determined by step  d)  iii)  must  agree
     exactly with the CF posted on the analyzer.   ASTM round-off
     shall be used.

     iv) The mean propane value determined  in step c) v) must agree
     with the propane calibration gas  within  1 percent of  the propan
     calibration gas concentration.

-------
                                    86
F.  Sample System Test Procedures

     1.  Sample Cell Pressure Variation, Low Flow,  and Response Time

          a) Test Equipment

               i) Candidate instrument system.

             -  ii) One span gas cylinder with a concentration  between  70  and
               90 percent of full scale, and bottled  zero  gas.

               iii) One gauge (0-30 inches of water).

               iv) One pinch clamp.

               v) One Tedlar sample bag  (approximately 3 cubic foot capacity)

               vi) One 3-way ball valve, minimum  orifice size  is  .180  inches.

               vii) One stop watch or timer.

               ix) Associated fittings and lines  (non-reactive material).

          b) Testing Sequence

          This  test  procedure  is  written around parallel HC and CO  analy-
          zers, and  should  be performed on each  analyzer.   If the analyzers
          are  in  a  series  configuration  then  an additional  pressure  gauge
          will be  required,  but the test  can be  performed  on  both analyzers
          at the same time.

               i) Identify the  sample line entering the  sample cell.

               ii)  Install a  tee  fitting in  the  sample line  immediately
               upstream  of  the  sample  cell  (in  as close  as  practical).
               Install the tee  with the  branch pointing  up.

               iii) Connect the gauge to the  tee  with the  suitable  length of
               flexible tubing.

               iv) Warm up the  analyzer.

               v) Zero the analyzer.

               vi) Span  the  analyzer with  the span gas  through the  spanning
               network.   Record  the  pressure  gauge reading,  and  the  span
               response.

               vii)  Fecheck  the zero.    If  the zero has  shifted repeat steps
               v), vi), and vii).

               viii)  Switch  the analyzer  from span gas  flow  to  sample  flow
               (pump on).

-------
                     87
ix) Record  the  maximum and minimum  fluctuations  in the gauge
reading.

x) Attach the pinch clamp  to  the pressure  gauge  line.   Slowly
pinch  the  line until  the gauge  fluctuations are  minimized.

xi) The gauge reading should  be approximately halfway  between
the  readings recorded  in  step  ix) .   Record the  stabilized
gauge reading in step x) .

xii) Attach  the common port  of  the  3-way valve  to the probe
with a leak  free adapter.

xiii) Attach a  short length  of  tube to one  of  the remaining
two ports of the three-way valve, and  sample  room air  through
the tube.

xiv) Fill the sample bag with span gas.

xv) Attach  the  sample bag to the remaining port  of the 3-way
valve.

xvi) The 3— way  valve should be sampling  room  air.

xvii) switch the  3-way valve  to  the sample bag,  compare  the
stabilized concentration reading of  the  bag to the reading in
step vi) .   A difference of more than  1  percent  from the con-
centration  reading  in step vi) could  indicate a  leak  in  the
system.  Repair any leaks, and restart  the procedure  at step
xviii) Recheck  the zero by  switching  the  3  way  valve back to
room air.

xix) Identify the  concentration  value  corresponding to 95% of
the span gas in the bag.  Record  the 95% value.

xx) Switch  the  3-way valve  to  the sample bag and  simultane-
ously start a timer.

xxi) When  the  analyzer response  reaches  the 95%  value  stop
the timer.  Record the  elapsed time.

xxii)  Record  the  sample  cell  pressure   after  the  analyzer
reading  has  stabilized.    Record  the  stabilized  analyzer
concentration reading.

xxiii) Identify the concentration value  (5%)  corresponding to
the  stabilized reading in  step  xxii) (100%)  minus  the  95%
value identified in step xix) .

xxiv)  switch  the  3-way  value  back  to room air  and  simul-
taneously start the timer.

-------
     xxv) When  the  analyzer reaches  the  5% value,  stop  the  timer.
     Record the elapsed time.

     xxvi)  Connect   the  needle  valve between  the  probe and  the
     3-way valve.

     xxvii) Adjust  the needle valve  until  the low  flow  indicator
     is just barely activated.

     xxviii) Repeat steps xvi) through xxv).

     xxix) Use  the  same  95% and 5% values identified  in  step  xix)
     and xxiii).

     xxx) Record the 5% and 95% response  times.

     xxxi) Record the stabilized concentration  reading, and  sample
     cell pressure in the same manner used in  step  xxii).

c) Calculations

     i)  Compute the difference  in pressure  between step vi)  and
     step xi),  and  step  vi)  and  step xxii).   The values will  be
     the  pressure   difference  between  spanning   and   sampling.

     ii) Compute  the  difference  between  maximum  and minimum pres-
     sure readings in step b)  ix).  The value will  be  the pressure
     variation during sampling.

     iii) Compute the  difference  in  pressure between  step xi)  and
     step  b)   xxxi).   The  value-  is  the difference  in  pressure
     variation  between  normal  flow  and  low  flow  conditions.

     iv) Compute  the percentage  change  in  analyzer gas  response
     between step b) xxii) and step b) xxxi).

d) Acceptance Criteria

     i)  If  the  calculated value in steps c)i,  and  c)iii  are  less
     than 4 inches  of  water,   and  less  then  6  inches  of  water  for
     step  c)ii,  the  pressure variations  in  the  analyzer  flow
     system under these test conditions are acceptable.

     ii) If the change in analyzer response computed  in  step  c)iv
     is  less  than  1.5  percent and the  elapsed times recorded  in
     step b) xxx) are less than 12 seconds, the low flow  indicator
     system  is acceptable  as well as  the system response  time.

     iii) If the elapsed time  in steps b) xxi)  and  b) xxv) is  less
     then 12 seconds, the system response time  is  acceptable under
     normal conditions.

-------
                     89
iv) It  is  recommended that the  sample  cell pressure  gauge(s)
remain hooked  up  for the duration of the analyzer  check  out.
If the  sample  pressure varies by more  then 4  inches  of water
from  the  most recent  span pressure,  then  the acceptance  in
d)i is void.

-------
2.  Maximum Sample Cell Pressure Variation during Sampling Test Procedure

    (See Test Procedure F.I)

3.  Maximum Sample Cell Pressure Variation Between Normal Flow
    and Low Flow Indication Test Procedure

    (See Test Procedure F.I)

4.  Response Time Test Procedure

    (See Test Procedure F.I)

-------
                               91
5.  System Leakage Test Procedures

     a) Equipment Required

          i) Candidate instrument.

          ii) One tee fitting.

          iii) One needle valve.

          iv) One span gas 70 to 90 percent of  full  scale  on  the  low
          range, and bottled zero gas.

          v) Associate lines and fittings  (non-reactive).

     b) Test Sequence

     This sequence  is written  assuming  that  the analyzer  incorporates
     the automatic leak checking feature specified.

          i) Install  the  tee fitting between  the  probe and  the  sample
          line.

          ii) Attach  the  needle valve to  the remaining  port  of  the  tee
          and close the valve.

          iii) Warm up the analyzer.

          iv) Zero and span  the analyzer with the analytical  gases.

          v) Record the span value.

          vi) Place the probe in the  leak  check  receptacle.

          vii) Record the analyzer response to span  gas  introduced
          through the leak check recepticle, probe,  and  sample
          line.

          viii) If  the  analyzer response  in step vii) differs from  the
          response  in step  v)  by  more  than  1 percent  of  the value
          recorded  in step  v),  check  the  system  for  leaks or other
          problems.  Restart the procedure at step iii).

          ix) Gradually open the needle valve until  the  leak  check  fail
          light just comes on.

          x) Record the analyzer response.

          xi) Use  the bubble  check method  to   check  for  leaks   in  the
          spanning system.

-------
                          92
c) Calculations

     i) Compute  the percent  leakage by  subtracting  the  response
     recorded  in  step  b) x) from  the  response in step  b)  v),  and
     then  dividing  result  by  the value  recorded in  step b)  v).
     Multiply  the result by 100.

d) Acceptance  Criteria

     i) If  the value calculated  in  step  c)  i)  is  equal  to or less
     than  the  leakage  rate   specification   (Chapter  IV,   Section
     F.5.),  the system  leakage  indication  system is acceptable.

     ii)  If  the spanning  system  shows  significant  leaks  at  fit-
     tings, attempt  to repair  the  leaks by tightening  or replacing
     the  fittings.    If   leaks occur  in other locations  of  the
     spanning  system, or appear  to be  the result  of  system design.
     an  engineering  report  must be  submitted  by  the  analyzer
     manufacturer  describing  the causes and preventive  remedies
     for the leak prior  final  acceptance  of  the system.

-------
                                    93
G.  Operating Environmental Test Procedure

     1.  Test Procedure

          a) Equipment required

               i) Candidate instrument

               ii) Environmental chamber with  temperature capability between
               35°F  and  110°F,  and humidity  capability between  10  and 9P
               percent relative humidity.

          b) Test Conditions

               i) 105°F (+ 5%°F) with a relative humidity between  96 and  100
               percent (non-condensing).

               ii) 40°F  (+  5°F)  with a relative  humidity  between 75 and 80
               percent with a 10 mph wind.

               iii)  35°F  (+  5°F)  with a relative humidity between 15 and 25
               percent.

          c) Required Tests of each  test condition  (Test Procedure numbers
          are given  in parentheses).

               i) System Warm-up (H.I.a))

               ii) System Leakage  (F.5.)

               iii)  Sample Line Crush Test '(C.2.)

               iv) Analyzer Calibration Curve  (E.I.)

               v) Pressure and Temperature Compensation  (E.3. b) and c))

               vi) Analyzer Zero and Span Drift (E.4.)

               vii)  Sample Cell Temperature  (as applicable, see E.7. and H.I.)

               viii) Analyzer Water  Interference  (E.8.)

               ix) Sample Cell Pressure Variation, Low Flow, and Response
               Time  (F.I.)

               x) HC hang-up (C.4)

               xi) If used, automatic zero/span system (H.2. a))

               xii)  If used, automatic read  system  (H.2. b) i) 1))

               xiii) If used, anti-dilution  system

-------
     xiv) If used, printer

     xv) If used, automatic data collection system

d) Test Sequence

     i) The  analyzer  shall be turned off and allowed  to  stabilize
     at least 3 hours at the test condition prior to beginning the
     performance check.

     ii) The warm-up test followed by the leak check test  shall be
     the first two tests performed at each test condition.

     iii) The  remaining tests may be  performed  in any convenient
     order.

     iv) An  additional  leak check shall be  performed  at  the com-
     pletion  of  the  required  tests  for   each   test condition.

     v)  The  first   test  condition  shall  be  the  35°F  condition
     followed by the 40°F and the 105°F conditions.

e) Acceptance Criteria

     i) If  the  analysis system passes  all  of  the individual test
     requirements as specified by each  test in c) and  d),  then the
     environmental  operating  characteristics  of  the  system are
     acceptable.

-------
                                    95
H.  Fail-Safe System(s)

     1.  Test Procedures for Features Required  for All Systems

          a) Warm-up Lock-out Test Procedure

               i) Equipment Required

                    1) Candidate analyzer.

                    2) The equipment required  to  perform  the Analyzer  Sample
                    Cell Temperature Test Procedure  (see  Section E.7.).

                    3) The equipment required  to  Perform  the Analyzer  Zero
                    and Span Drift Test Procedure (see Section  E.4.).

                    4) A timer.

               ii) Test Sequence

                    1) Follow the basic measurement  preparations indicated
                    in Test Procedures E.4. and E.7.

                    2) After the analyzer has  stabilized  at  the ambient
                    conditions as determined by Test Procedure  E.7.  Turn on
                    the analyzer power and  simultaneously start a  timer.  (A
                    chart recorder may be used).

                    3) The chart recorder must  indicate both negative  and
                    positive zero drift.  If necessary, offset  the  chart
                    recorder zero 5 units up scale.

                    4) As soon as the warm-up  lock-out feature  deactivates,
                    record the elapsed time from  power on,  the  sample  cell
                    temperature, and immediately  zero and span  the  analyzer
                    with analytical gases per  manufacturer's operating
                    instructions.  Simultaneously, start  the chart  recorder
                    (in already started, mark  the chart)  at  a minimum  chart
                    speed of 0.5 inches per minute.

                    5) Immediately after spanning, begin  sampling  ambient
                    air (or analytical zero gas at the ambient  temperature)
                    through the sample line.

                    6) Monitor the zero drift  and the sample cell  temperature
                    for 5 minutes.

                    7) Record the lowest sample cell temperature during  the
                    5 minute period.

-------
                     96
iii) Calculations

     1) Locate on the chart the maximum and minimum  analyzer
     zero-response during the 5 minute period.

     2) Compute the difference  (as a percentage  of full  scale
     chart deflection) between  the analyzer zero-response
     determined by the span check in step b) 4)  and  the
     maximum zero-response, and then the minimum zero-response
     as identified in step iii) 1).

     3) Determine the largest difference in step c)  2) and
     multiply it by 2 	 (zero drift).

iv) Acceptance Criteria

     1) If the temperature observed in step ii)  2) is  approx-
     imately the same as the ambient temperature, the  analyzer
     is properly stabilized.

     2) If the temperatures recorded in step ii)  4)  and  step
     ii) 7) are equal to or greater than the specifications
     for sample cell temperature, the sample cell temperature
     is acceptable.

     3) While monitoring the sample cell temperature during
     the analyzer check-out procedure, if the  temperature  is
     less than the specifications when concentration data
     (span or sample) is read,  the acceptance  in step  2)  is
     void.

     4) If the zero drift, as calculated in step iii)  3)  is
     less than the specifications for zero drift (Chapter  IV,
     Section E.), then the zero drift after warm up  is
     acceptable.

     5) Acceptance of the above criteria constitutes accep-
     tance of the Warm-up Lock-out system.

     6) If the analyzer manufacturer indicates a typical
     lock-out elapsed time to the ultimate user,  then  the
     manufacturer must show that the elapsed time supplied
     to the user is truly typical if the time  recorded in
     step ii) 4) is more than 30 percent longer  than the
     time indicated to the user.

-------
                          97
b) Analyzer Low Flow Test Procedure

     i) See Test Procedure F.I., Sample Cell Pressure Variation,
     Low Flow, and Response Time.

     ii) Visual Observation of Features

c) Analyzer Leak Check Test Procedure

     i) See Test Procedure F.5., System Leakage Test Procedure

     ii) Visual observation of features

     iii) Self explanatory test procedure for timer check-out.

-------
                               98
2.   Test Procedures for Features Required on Decentralized Systems

     a) Automatic Zero/Span Check Test Procedure

          1) If the analysis system is equipped with the automatic gas
          span feature, the gas spanning feature must be used throughout
          the entire evaluation procedure unless noted otherwise (e.g.
          pressure and temperature compensation, Section E.3. b)).

          ii) The automatic span system shall be used with calibration
          gas (instead of span gas) for the Calibration Curve Test
          Procedure (E.I.).

          iii) Self-explanatory Test Procedures for the other require-
          ments.

     b) Automatic Read Feature Test Procedure

          i) If the analysis system is equipped with the automatic
          read feature, the system shall be deactivated for all
          evaluation testing except:

               1) The response time portion of  test procedure F.I.
               shall be rerun with the automatic read feature opera-
               tional.

               2) All system correlation tests  (Procedures in Section I)
               shall be run with the automatic  read feature operational.

               3) Other tests self explanatory.

     c) Printer Feature Test Procedure

     If the analysis system is equipped with a  printer, the printer
     shall be checked for proper operation.  The printer shall provide
     the official system results to be used in  the correlation proce-
     dures (Section I).

     d) Vehicle Diagnosis Feature Test Procedure

     Check for proper operation.  No other test procedures are
     required.

     e) Anti-Tampering Feature Test Procedure

          i) Visual observation

-------
                                    99
I.   System Correlation Test Procedures

The correlation procedure is designed to evaluate the performance of  the I/M
analyzers when sampling  auto exhaust.  The NDIR  procedure is slightly dif-
ferent from  the  FID HC correlation procedure.   However,  it is expected the
data  collection   phase  of  each  procedure can  occur  simultaneously.   For
clarity they will be presented as separate procedures.

     1.  NDIR Analyzer Correlation Test Procedure

          a) Equipment Required

               i) Candidate instrument.

               ii) A raw exhaust CO analysis system meeting  the requirements
               of 40CFR 86, Subpart D, for gasoline-fueled engines.

               iii) A  laboratory  grade NDIR HC analyzer  substantially  simi-
               lar  in  quality to  the laboratory  CO  analyzer, and operated
               according  to  the  requirements  for general  NDIR analysis of
               gasoline-fueled engines  contained  in  40  CFR 86,  Subpart D.

               iv)  A  tailpipe extension  that meets  the  probe location re-
               quirements in 40CFR 86 Section 312-79  (c)  (v).

               v) Two test vehicles.

               vi)  Calibration gases  for each range  used on both the candi-
               date  analyzers and  the  reference analyzer.  In  this  test
               procedure,  calibration gase.s  will  be used  instead  of span
               gases  to  span both analysis  systems.   The calibration  gases
               should  be between 70  and 90  percent  of  full  scale  on each
               range  used.    Calibration  gases  for  the  Subpart  D analysis
               system  shall   meet  or  exceed  the  requirements specified in
               Subpart D.

               vii) Option:   An adjustable dilution  box  and  mixing  chamber
               may  be  used  to obtain different exhaust concentration levels
               from the  vehicle.   The dilution box and mixing chamber  would
               be  installed   between  the  tailpipe and the  analysis system
               probes.

               viii)  Option:  A  chassis  dynamometer would  be  useful for
               loading the vehicle  to obtain different emission levels, but
               is not  required.

          b) Test vehicles

               i)  One 1975 or later  non-catalyst light-duty  vehicle (LDV),
               LOT, or HDG is acceptable.

               ii)  One  1978  or  later oxidation  catalyst  equipped  vehicle
               with air  injection.

-------
                           100
     ill) All  vehicles must  be  in  good operating condition  with
     all  emission control  systems  functional.   A  1975 FTP  type
     test  may  be  performed on  the vehicle  to  verify  functional
     operation of the emission controls.

c) Test Procedure Overview

The correlation procedure consists of  testing  the candidate  analy-   a^^
zer  at several  points on  each  range.   The  range  scale  of  the   ^^^
candidate  analyzer determines  the approximate  test  points.   The   5JS
exact  test  points  are then determined  by  the  concentration  levels   *^j^*i
observed by  the  reference system.  The exact  test points are  then   iLt^m
replicated  several  times  (minimum of  6)  based  on  the reference   £MB£
system  response  values.   This replicate data  is then analyzed  to
determine correlation  between  the candidate system and  the  refer-
ence system.

d) Test Sequence

     i)  Select  a test  vehicle,  and warm  up  that vehicle with  at
     least  30 minutes  of  hard  load.   (freeway  operation,  hard
     accelerations, etc.

     ii) Prepare  the  analysis  systems  for measurement  (i.e. warm-
     up, spanning etc.)  operate the reference system according  to
     the provisions in Subpart D where  applicable.

     iii) Insert  the probe  of the  candidate  instrument into  the
     tailpipe approximately 16 inches.

     iv)  Operate  the vehicle (or  adjust  the  dilution  box)  to
     obtain stable concentration readings  at approximately 20,  50,
     70  and  90 percent  of  full  scale  concentration  value  on  the
     low range of the candidate analyzer.

     v)  Select  the lowest  useable range  on  the Subpart D  system
     for each test point in d)iv)  (see  § 86.338).

     vi)  Sample  for  approximately 1  minute at  each  test  point.

     vii)  Record  the average emission  value  of  the candidate    j^^^
     instrument and  the average  chart  deflection from  the  refer-    _~Sr~
     ence system  over the  last  10  seconds  for  each  test  point.    "™^t
     The  two  readings recorded  at each test  point  should be  re-    CSS
     corded  over  the same  time  frame,  and  are  defined   as  a    CMMS
     "paired" data point.

     viii)  Select  the  high  range of   the  candidate  instrument.

     ix)  Operate  the vehicle (or  adjust  the  dilution  box)  to
     obtain  stable  readings at  approximately 20,  50,  70,   and  90
     percent  of  full  scale on  the  high  range  of  the candidate

-------
                          mi
     instrument.  If  the  higher emission values require  impracti-
     cal  vehicle  operation,  cease  data generation  and continue
     with the test procedure.  At least  one vehicle should be able
     to achieve the higher values.

     x)  For  the  attainable  test  points  in  the  preceding step,
     repeat step d)vii).

     xi)  The chart  deflections recorded  in  steps  d)vii) and d)x)
     for  the  Subpart D  system are  now  defined  as "reference set
     points" that will  be used for  the  remainder  of the correla-
     tion  test.   When repeating  the test  sequence  with the same
     vehicle or a different vehicle, the  vehicle  or dilution box
     should be  adjusted to obtain, as close as possible,  the exact
     "reference"  chart  deflection.   For the  test  points not at-
     tainable in step d)ix), the chart deflection of the  reference
     system becomes  the  set  point  the  first  time  a data pair  is
     recorded for that test point.

     xii) Repeat  steps  d) iv) through d) x) at the set  points de-
     scribed in step d)xi).  Record  the emission  values from the
     candidate   system,  and the chart deflections  from  the  refer-
     ence system.

     xiii) Ground or short two plug  wires on V8 engines  (opposite
     sides of  intake manifold),  or  one  plug  wire  for 6 cylinder
     and 4 cylinder engines.

     xiv) Perform steps  d)iv)  through d)x) at  the  set points de-
     scribed in step  d)xi) a minimum of three times.   Record the
     emission  values  from the  candidate  system,  and  the chart
     deflections from the reference  system.

     xv)  Select  the  2nd test  vehicle,  and  repeat  steps d)iv)
     through d)xiv)  at  the set  points  described  in  step  d)xi).
     Record the emission  values from  the candidate system, and the
     chart deflections from the reference system.

     xvi) Repeat the procedure as necessary to obtain a  minimum  of
     6 replicate responses at each test  point.

e) Calculations

     i) For the candidate  system compute  the mean (x) and standard
     deviation(s)   of  the emission  values  for  each  test   point.

     ii) For the reference system convert each chart deflection  to
     a  concentration value  from  the analyzer  calibration  curve.

     iii) For  the  reference system  compute the mean (x)  and stan-
     dard  deviation(s)  concentration values  for  each  test  point.

-------
                          102
     iv)   Compute  the  normalized  precision difference  ( A P) for
     each test point by:
          A?

     where:
((Ks/ x)  candidate)  - ((Ks/ x) reference)
     Sample Size

          5
          6
          7
          8
          9
           JC

          3.5
          3.1
          2.9
          2.7
          2.6
Sample Size

    10
    11
    12
    13
    14
2.5
  46
  40
  36
2.31
     v)  For each  range of the candidate analyzer perform a  linear
     regression on  all of  the paired  data  that were measured on
     that  range.   Force the  regression through zero.  The  refer-
     ence  system  is  the  independent  variable.    Identify   the
     slope(m)  of the regression line.

     vi)  For  each  range  of   the  candidate  analyzer compute  the
     ratio(R)  of  the  analyzer mean concentrations  for each  test
     point corrected  for  the  slope(m)  identified in  the preceding
     step by:

          R = (x reference/x candidate)m'

f)  Acceptance Criteria

     i)  Identify  the  largest  A P value.   If  the  largest  A? value
     is  less  than  or  equal to the specification  for  A  P. then  the
     in-use  precision  of  the  candidate  system  is  acceptable.

     ii) If the  slope (m)  for  each range of the  candidate analyzer
     is  within  the  limits for slope,  then the slope test results
     are acceptable.

     iii)  Identify  the minimum and  the maximum ratio(R) of slope
     corrected  mean  concentration  values.    If the  minimum   and
     maximum  ratios are  within the  range  specified,  then the  mean
     concentration ratio test  results are  acceptable.

-------
                               103
2.  FID Analyzer Correlation Test Procedure

     a) Equipment required

          i) Candidate instrument.

          ii) An HC analysis system meeting the requirements of 40CFR86,
          Subpart D for gasoline-fueled engines.

          iii)  A  tailpipe  extension   that  meets  the  probe  location
          requirements in 40CFR86 Section 312(c)(v).

          iv) Three test vehicles.

          v) Calibration  gases  for each range  used  by both the candi-
          date  analyzers  and  the  reference  analyzer.   In  this  test
          procedure,  calibration  gases  will  be  used  instead  of  span
          gases  to  span  both  analysis  systems.   The  gases  should  be
          between 60  and  90 percent of  full  scale  on each range used.

          vi) Calibration gases for the Subpart D analysis  system shall
          meet  or  exceed  the   requirements  specified  in  Subpart  D.

          vii) Option:   A dilution system that allows  the  auto exhaust
          to be  diluted  in a controlled manner with a suitable mixing
          chamber,  may  be used  to  obtain  different exhaust concentra-
          tion levels from the vehicle.

          viii)  Option:    A  chassis dynamometer  may  be  used  with  or
          without the dilution system to obtain different concentration
          levels.

     b) Test Vehicles

          i) The  type of test vehicle required for  the FID correlation
          test are  the  same as the type required for  the NDIR correla-
          tion.   Generally it is  preferred  that  the same vehicles be
          used for both correlations.

     c) Test Procedure Overview

     The correlation  procedure  consists of testing  the candidate anal-
     yzer at several  points on  each  range.    The  range  scale  of  the
     candidate  analyzer determines  the approximate  test  points.   The
     exact  test  points  are then determined by  the concentration levels
     observed by  the  reference system.  The exact  test points are then
     replicated  several  times  (minimum of 6)  based  on  the reference
     system  response  values.   This replicate data  is then analyzed to
     determine correlation  between the candidate system and the refer-
     ence system.

-------
                           104
d) Test Sequence

The  test  sequence  is identical  to CO  correlation test  required
with the following exception.

     i) Stabilized concentration readings  of  approximately 10,  20,
     35, 50, 70, and 90 percent of  full  scale concentration values
     should  be  substituted  for  the  value  in step  l.d)iv)  and
     l.d)ix).   Record  the HC  concentration value  from the candi-
     date analyzer as ppmh or  ppmC6.

e) Calculations

     i) For the candidate system compute the  mean  (x)  and  standard
     deviation(s)  of  the  emission  values  for each  test point.

     ii) For the reference system convert  each chart deflection to
     a  concentration value  in ppm  C3  (propane) from  the  analyzer
     calibration curve.

     iii)  For  the  reference system  compute the mean  (x) and  stan-
     dard deviation(s)  concentration  values  for each test point.

     iv)  Compute  the  normalized  precision  difference  (  AP)  for
     each test point by:

           A P = ((Ks/ x) candidate) - ((Ks/  x)  reference  FID)

     where:

     Sample Size          JC          Sample Size        _K

          5              3.5             10            2.5
          6              3.1             11            2.46
          7              2.9             12            2.40
          8              2.7             13            2.36
          9              2.6             14            2.31

     v) For  each range of the candidate analyzer  perofrm  a linear
     regression  on all of  the paired data that were measured  on
     that  range.   Force the  regression through zero.  The refer-
     ence  system  is  the  independent   variable.   Identify  the
     slope(m) of the regression line.

     vi)  For each  range of   the  candidate  analyzer  compute  the
     ratio(R)  of the  analyzer mean concentrations  for each test
     point corrected for the slope(m)  identified in the preceeding
     step by:

          R = (x reference FID/x candidate)m

f) Acceptance Criteria

-------
                     105
i) For HC analyzers manufactured before January  1986,  the
following data shall be available:

     1) Precision
     2) Slope Comparison
     3) Ratio of Modal Averages

ii) For HC analyzers manufacturered after January  1986,  the
analyzer shall meet the following criteria:

     1) Identify the largest  A p value.  If the  largest  /\ P
     value is less than or equal to the specification  for  A
     then the in-use precision of the candidate  system is
     acceptable.

     2) If the slope(m) for each range of the candidate
     analyzer is within the limits for slope, then  the slope
     test results are acceptable.

     3) Identify the minimum and the maximura ratio(R)  of
     slope corrected mean concentration values.  If  the
     minimum and maximum ratios are within the range speci-
     fied, then the mean concentration ratio test  results
     are acceptable.

-------