SAMPLING AND ANALYSIS
        PROCEDURES FOR
          SCREENING OF
    INDUSTRIAL EFFLUENTS
FOR PRIORITY POLLUTANTS
   UNITED STATES ENVIRONMENTAL PROTECTION AGENCY
          EFFLUENT GUIDELINES DIVISION

             WASHINGTON, D.C.


           MARCH 1977 revised APRIL 1977

-------
                                FOREWORD
     These guidelines for sampling and analysis of industrial wastes
have been prepared by the staff of the Environmental Monitoring and
Support Laboratory, at the request of the Effluent Guidelines Division,
                                               •
Office of Water and Hazardous Wastes, and with the cooperation of the
Environmental Research Laboratory, Athens, Georgia.  The procedures
represent the current state-of-the-art but improvements are anticipated
as more experience with a wide variety of industrial  wastes is obtained.
Users of these methods are encouraged to identify problems encountered and
assist in updating the test procedures by contacting  the Environmental
Monitoring and Support Laboratory, EPA,  Cincinnati, Ohio 45258.

-------
                              CONTENTS




ORGANICS BY PURGE AND TRAP -  GAS CHROMATOGRAPHY	    1




     Scope	    1


     Special Apparatus and Materials   .  .  .  ,	    2


     Gas Chromatographic Column Materials  	    2


     Procedure   	    3
                                                 X"

          Preparation of Standards  	    3


          Preliminary Treatment of Sample  	    4
                                             «

          Purging and Trapping Procedure   	    4


          GC-MS Determination 	    6


               Purge Parameters	    6


               Gas Chromatographic Parameters  	    7


               Mass Spectrometer Parameters  	    7


          Quality Assurance 	    8


               Precision	    9


               Calibration of GC-MS System   	  10


          Qualitative and Quantitative Determination   ....  10


     Reporting of Data	11


     Direct Aqueous Injection Gas Chromatography  	  11




ORGANICS BY LIQUID-LIQUID EXTRACTION - GAS CHROMATOGRAPHY . .  16




     Scope	16


     Special Apparatus and Materials  	  16


     Procedure	16


          Base-Neutral Extraction 	  17

-------
               Separatory Funnel Extraction 	  17



          Acid (Phenols)  Extraction	  18



          Emulsions	19



               Continuous Extraction  	 .......  20



          Blank Extraction	20
                        s'


          Pesticides	21



     GC-MS Analysis	23

                •

          Base-Neutral  	  23



          Acid	25



     Quality Assurance  	  29



     Reporting of Data	30





METALS	43




     Sample Preparation ..... 	  43



     Apparatus	44



     Procedure	44



     Quality Assurance  ... 	  47



     Data Reporting	4,8




CYANIDES	  49
                                                               »


     Sample Preparation 	  	  49

                                                               t

     Sample Procedure	49



     Quality Assurance  	  49



     Reporting of Data	49

-------
PHENOLS	50



     Sample Preparation  	 50



     Procedure	50



     Quality Assurance   ....  	 50



     Reporting of Data	50






REFERENCES	51






APPENDIX I	53



APPENDIX II	55



APPENDIX III	63



APPENDIX IV	70

-------
                          LIST OF TABLES





Table I.  Elution Order of Volatile Priority Pollutants ...  12



Table II.  Characteristic Ions of Volatile Organics	14



Table III.  Pesticides	33



Table IV.  Base-Neutral Extractables  	  34



Table V.  Acid Extractables	36



Table VI.  Elution Order of Most of the Semivolatile



           Priority Pollutants  	  37



Table VII.  Order of Elution for OV-17 SCOT Column	41



Table VIII.  Metals	45

-------
                             Appendices




  I.   General Information



 II.   Possible Sources for Some Priority Pollutant Standards



III.   Collection of Samples for Screening Analyses

-------
                  UNITED STATES ENVIRONMENTAL PROTECTION AGENCY
SUBJECT:  Sampling and Analysis Procedures for Screening
          of Industrial Effluents for Priority Pollutants
 FROM:     William A. Telliard, Chief
          Energy and Mining Branch
 TO:       Project Officers
          Effluent Guidelines Division

THRU:     Robert B. Schaffer, Director,
          Effluent Guidelines Divisi
DATE:
                                                                     MAY 2 7 1977
          As you know, in the s-ettlement of several  cases in the District
          Court for the District of Columbia, the Environmental  Protection
          Agency has agreed to review and revise regulations based on the
          Best Available Technology Economically Achievable (BAT), New
          Source Performance Standards,  and Pretreatment Standards for 21
          industrial categories.

          In this revision, consideration is to be given to the  application
          of limitations of a list of 65 materials appearing in  Appendix A
          of the Settlement Agreement.   These materials  are generally
          referred to as priority pollutants.  The priority pollutants are
          both single compounds and families of compounds.   The  Agency has
          established an unambiguous list of 129 compounds  which it
          believes fulfills the requirements of the  court order  and can be
          analytically determined.
          To maintain consistent sampling and analytical  procedures  the
          Agency has developed a sampling protocol  and analytical  methods
          to be used for screening for priority pollutants.   This  protocol
          represents the most current procedures for the  sampling  and
          analysis of these priority pollutants.  Because of the large
          number of analysis required, argon plasma atomic-emission
          spectroscopy will be used by the Agency for most metals  analysis,
          Pertinent information about this analytical  method, which  is  an
          accepted alternate method under section 304(g)  is  attached.

          The data gathering process basic to revising the regulations
          consists of two phases.   The initial  phase is the  screening
          sampling and analysis procedure to ascertain the presence  or  •
EPA Form 1320-6 (Rev. 6-72)

-------
absence of the priority pollutants.  The second phase or
verification sampling will be used to quantify those pollutants
found to be present during the screen sampling.

These materials are made available for your information and use
during the screening phase only.

-------
                  UNITED t,. ATES ENVIRONMENTAL PROTECTIL.. AGENCY


SUBJECT: Approval of Alternate Test Procedure - Automated      DATE:   . .  »• O 9 19/6
         Simultaneous Analysis of  Twenty Selected  Elements
         by Inductively Coupled Argon Plasma Emission Spectrotcopy
FROM:    Mr. Francis T. Mayo
         Regional Administrator, Region V

TO:      Mr. Thomas E. Yeates, Director
         Central Regional  Laboratory,  Region V

THRU:    Mr. Chris Timm, Director   '                /
         Surveillance & Analysis Division, Region ^/


         The Environmental Monitoring  and Support Laboratory (EMSL)-- Cincinnati,
         EPA has carefully reviewed your application  for  use of an alternate test
         procedure for the automated simultaneous analysis of twenty  elements by
         emission spectroscopy using the inductively  coupled argon plasma  as the
         emission source.  Your application  specifies that the automated instrumen-
         tal methodology will  be used  at the Central  Regional Laboratory,  Region V
         for all sample types  applicable to  the National  Pollutant Discharge Elimi-
         nation System (NPDES).
                                                        >
         The proposed method uses  the  sample digestion procedure of 40 CFR Part  136,
         but instead of utilizing  referee atomic absorption  spectrometry,  the
         digested sample is aspirated  into a  high temperature inductively  coupled
         argon plasma (ICAP),  and  several total elemental concentrations in  the
         aspirated sample  are  measured  simultaneously using  an emission spectro-
         graph with an appropriate photomultiplier tube for  each element.  The
         calculation of the elemental  concentrations  is done by a  computer inter-
         faced to the spectrograph.  Twenty  elements, primarily metals, have been
         selected by the Central Regional Laboratory  for analysis  by ICAP-emission
         spectroscopy.   In addition to  total  elemental concentrations, the proposed
         methodology can readily measure dissolved concentrations  for the  same
         elements simply by filtering  a  suitable aliquot  through a  0.45 y membrane,
         acidifying the filtrate as necessary for preservation, eliminating  the
         digestion procedure and aspirating  the filtrate into the  ICAP-emission
         spectrograph.

         The comparability data  you have provided for ICAP-emission spectroscopy
         and referee atomic absorption spectrophotometry indicate  the two methods
         yield equivalent  data for  a variety of waste effluents representative of
         the NPDES.   ICAP-emission  spectroscopy is shown to provide a comparable
         or  superior performance, depending on the element,  for the measurement of
         recovery and  precision  for random element "spikes"  of  NPDES waste effluents.
         Although the  stated detection lira-its and lowest quantitatively deterroinable
         concentration vary slightly from day-to-day and are a  function of the ICAP
         nebulizer,  the  reportable detection limits for ICAP-emission spectroscopy
EPA PM» 1320-4 (ft.*.

-------
                    Organics by Purge and Trap -



                        Gas Chromatography





1.   Scope



          This method is designed to determine those "unambiguous



     priority pollutants," associated with the Consent Decree, that



     are amenable to the purge and trap method   .   These compounds



     are listed in Table I of this section.   It is  a gas chromato-



     graphic-mass spectrometric (GC-MS)  method intended for quali-



     tative and semi-quantitative determination of  these compounds



     during the survey phase of the industrial effluent study.




          Certain compounds, acrolein and acrylonitrile, are not



     efficiently recovered by this method and should be determined



     by direct aqueous injection GC-MS.   Direct aqueous injection



     GC-MS is recommended for all compounds  that exceed 1000 ug/1.




          The purge and trap and the liquid-liquid  extraction methods



     are complementary to one another.   There is an area of overlap



     between the two and some compounds  may  be recovered by either



     method.  However, the efficiency of recovery depends on the



     vapor pressure and water solubility of  the compounds involved.



     Generally, the area of overlap may  be identified by compounds



     boiling between 130°C and 150°C with a  water solubility of



     approximately two percent.   When compounds are efficiently re-



     covered by both methods,  the chromatography determined the



     method of choice.  The gas  chromatographic conditions selected

-------
                                 - 2  -

     for the purge and trap method are, generally, not  suitable  for

     the determination of compounds eluting later than  chlorobenzene


2.   Special Apparatus and Materials


          Sample extraction apparatus  (minimum requirements):

               5-ml glass syringes with Luer-Lok - 3 each

               2-way syringe valves  (Teflon or Kel-F) - 3 each

               8-inch, 20 gauge syringe needle - 2 each

               5-ml glass, gas-tight  syringe, pressure-lok

                    or equivalent - 1 each

               Tekmar Liquid Sample Concentrator, model LSC-1

                    or equivalent.  Includes a sorbent  trap

                    consisting of 1/8 in. O.D.  (0.09 to 0.105

                    in.I.D.)        x6in.  long stainless steel tube

                    packed with 4 inches of Tenax-GC  (60/80 mesh

                    and 2 inches of Davison Type-15 silica gel

                    (35/60 mesh).


3.   Gas Chromatographic Column Materials


          Stainless steel tubing 1/8  in. O.D.  (0.09 to  0.105 in.

     I.D.) by 8 ft. long.  Carbopack  C  (60/80 mesh) coated with
                       (c)
     0.2% Carbowax 1500   .  Chromosorb-W (60-80 mesh)  coated with

     3% Carbowax 1500.
      (a)  Available from Precision Sampling Corp., P.O. Box 15119,
          Baton Rouge, LA  70815.

      (b)  Available from Tekmar Company, P.O. Box 37202,
          Cincinnati, OH  45222.

      (c)  Available from Supelco, Supelco Park, Beliefonte, PA
          16823.  Stock No. 1-1826.

-------
                                - 3 -




4.   Procedure





          Preparation of Standards - Prepare standard stock solu-



     tions (approximately 2 ug/yl)  by adding, from a 100 yl'syringe,



     1 to 2 drops of the 99+% pure reference standard to methanol



     (9.8 ml)  contained in a tared 10 ml volumetric flask  (weighed



     to nearest 0.1 mg).  Add the compound so that the two drops



     fall into the alcohol and do not contact the neck of the flask.



     Use the weight gain to calculate the concentration of the stand-



     ard.  Prepare gaseous standards,  i.e.,  vinyl chloride, in a



     similar manner using a 5 ml valved gas-tight syringe with a



     2 in. needle.  Fill the syringe (5.0 ml)  with the gaseous com-



     pound.  Weigh the 10 ml volumetric flask containing 9.8 ml of



     methyl alcohol to 0.1 mg.   Lower the syringe needle to about



     5 mm above the methyl alcohol meniscus.   Slowly inject the



     standard into the flask.   The gas  rapidly dissolves in the



     methyl alcohol.   Reweigh the  flask,  dilute to volume,  mix,



     tightly stopper,  and store in a freezer.   Such standards  are



     generally stable  for at least one  week when maintained, at less



     than 0°C.   Stack  standards of compounds  which boil  above  room



     temperature are generally stable  for at  least four  weeks  when



     stored at 4°C.





          [Safety Caution:   Because  of  the toxicity of most' organo-



     halides,  primary  dilutions must be prepared in a hood.  Fur-



     ther, it  is advisable to  use an approved  respirator when



     handling  high concentration of  such  materials.]

-------
                            -  4  -




     From the primary dilution  prepare a  secondary  dilution



mixture in methyl alcohol so  that 20.0 yl of the  standard,



diluted to 100.0 ml in organic  free water, will give a  stand-



ard which produces a response close to that of the  unknown.



Also prepare a complex test mixture at a  concentration  of



100 ng/yl containing each of  the compounds to be  determined.



Prepare a 20 yg/1 quality check sample from the 100 ng/ul



standard by dosing 20.0  ul  into 100.0 ml  of organic free water,





     Internal Standard Dosing Solution -  From stock standard



solutions prepared as above,  add a volume to give 1000  yg each



of bromochloromethane, 2-bromo-l-chloropropane, and 1,4-



dichlorobutane to 45 ml  of  organic free  (blank water) con-



tained in a 50 ml volumetric  flask, mix and dilute  to volume.



Prepare a fresh  internal standard on a weekly basis.  Dose



the internal standard mixture into every  sample and reference



standard analyzed.





     Preliminary Treatment  of Sample - Remove samples from



cold storage  (approximately an  hour prior to analysis)  and



bring to room temperature by  placing in a warm water bath



at 20-25°C.





     Purging and Trapping Procedure - Adjust the  helium purge



gas flow to 40 ml/min.   Set the Tekmar 2-way valve  to the



purge position and open  the purging device inlet.   Remove



the plungers from two 5-ml  syringes and attach a  closed 2-way



syringe valve to each.   Open  the sample bottle and  carefully

-------
                            - 5  -




pour the sample  into  one  of the syringes  until it overflows.



Replace the  syringe plunger and compress  the  sample.   Open



the syringe  valve and vent  any  residual air while carefully



adjusting the volume  to 5.0 ml.  Then  close the valve.   Fill



the second syringe in an  identical manner from the same



sample bottle.   Use the second  syringe for a  duplicate analy-



sis as needed.   Open  the  syringe valve and introduce  5.0  yl of



the internal standard mixture through  the valve bore, then



close the valve.  Attach  the 8-inch needle to  the syringe



valve and inject the  sample  into the purging device.  Seal



the purging device and purge the sample for 12  minutes.  The



purged organics  are sorbed  on the Tenax-silica  gel  trap at



room temperature (20-25°C).





     While the sample is being purged, cool the  gas chromato-



graphic column oven to near room temperature  (20-30°C).  To



do this, turn heater  off and open column  oven door.





     At,the completion of the 12-minute purge time, inject



the sample into  the gas chromatograph by  turning  the valve



to the desorb position.  Hold in this position  for  four min-



utes while rapidly heating the trap oven  to 180QC, then return



the valve to the purge position, close the GC column oven



door,  and rapidly heat the GC oven to 60°C.  Consider this



time zero and begin to collect retention data.  Hold at 60°C



for four minutes, then program at 8°/minute to 170°C and hold



until all compounds have eluted.  Begin collecting GC-MS

-------
                            -  6  -
GC-MS data as soon as the GC-MS vacuum system has stabilized
(<10~5 torr).
     While the sample is being chromatographed, flush the
purging device with two 5-ml volumes of organic free water.
Then bake out the trap  (vent to atmosphere) to minimize the
amount of water desorbed into the GC-MS system during the
succeeding injection step.  [Note:  If this bake out step
is omitted, the amount of water entering the GC-MS system
will progressively increase causing deterioration of and
potential shut down of the system.]

     GC-MS Determination - Suggested analytical conditions
for determination of the priority pollutants amenable to
purge and trap, using the Tekmar LSC-1 and the computerized
Finnigan 1015 GC-MS are given below.  Operating conditions
vary from one system to another; therefore, each analyst
must optimize the conditions for his equipment.

Purge Parameters
     Purge gas - Helium, high purity grade
     Purge time - 12 minutes
     Purge flow - 40 ml/min.
     Trap dimensions - 1/8 in. O.D. (0.09.to 0.105 in. I.D.)
          x 6 in. long
     Trap sorbent - Tenax-GC, 60/80 mesh (4 in.) plus Type 15
          silica gel, 35/60 mesh  (2 in.)
     Desorption flow - 20 ml/min.
     Desorption time - 4 min.
     Desorption temperature - 180°C

-------
                            - 7 -




Gas Chromatographic Parameters



     Column - Stainless  steel,  8  ft.  long x  1/8  in.  O.D.



           (0.09 to 0.105 in.  I.D.) packed with Carbopack  C



           (60/80 mesh) coated with 0.2% Carbowax 1500,  pre-



          ceded by a 1 ft. x  1/8  in.  O.D.  (0.09  to 0.105  in.



          I.D.) packed with Chromosorb-W coated  with  3%



          Carbowax 1500.



     Carrier gas - Helium at  33 ml/min.





     Oven temperature.- Room  temperature during  trap  desorp-



tion, then rapidly heat to 60°C, hold at 60°C for four min-



utes, then program to 170°C at 8°/minute.  Hold  at 170°C  for



12 minutes or until all compounds have eluted.





Mass Spectrometer Parameters



     Data system - System industries System 150



     Separator - glass jet



     Electron energy - 70 ev



     Emission current - 500 ua



     Ion energy - 6 volts



     Lens voltage - (-)IOO volts



     Extractor voltage - 8 volts



     Mass range - 20-27, 33-260 amu



     Integration time/amu - 17 milliseconds



     Samples/amu - 1





     Gas Chromatographic Column Conditioning Procedure; -



Attach the Carbowax 1500-Chromosorb end of the column to the



inlet system of the gas chromatograph.  Do not,  at this time,

-------
                            -  8  -




attach the column exit to the detector.  Adjust the helium



flow rate through the column to 33 ml/minute.  Allow the



column to flush with helium for ten minutes at room tempera-



ture, then program the oven from room temperature to 190°C at



4°C/rainute.  Maintain the oven at 190°C overnight  (16 hours).





     Handle the column with extreme care once it has been



conditioned because the Carbopack is fragile and easily frac-



tured.  Once fractured, active sites are exposed resulting in



poor peak geometry  (loss of theoretical plates).  Recondition-



ing, generally, revitalizes the analytical column.  Once



properly conditioned, the precolumn may be removed.  The re-



tention data1listed in Table I was collected with the pre-



column in the system.





     Quality Assurance - The analysis of blanks is most



important in the purge and trap technique since the purging



device and the trap can be contaminated by residues from



very concentrated samples or by vapors in the laboratory.  Pre-



pare blanks by filling a sample bottle with low-organic water



(blank water) that has been prepared by passing distilled



water through a pretested activated carbon column.  Blanks



should be sealed, stored at 4°C, and analyzed with each group



of  samples.





     After each sample analysis, thoroughly, flush the purg-



ing device with blank water and bake out the system.  Sub-



sequently, analyze a sample blank  (one that has been transported

-------
                            - 9 -




to and from the sampling  site) .   If positive  interferences



are noted, analyze a  fresh  laboratory  sample  of  blank water.



If positive interference  still occurs, repeat the  laboratory



blank analysis.  If interference  persists, dismantle  the



system, thoroughly, clean all parts that the  sample, purge



gas and carrier gas comes into contact with and  replace or



repack the sorbent trap and change purge and  carrier  gas.





     Precision - Determine the precision of the  method by



dosing blank .water with the compounds selected as  internal



standards - bromochloromethane, 2-bromo-l-chloropropane, and



1,4-dichlorobutane - and running replicate analyses.  These



compounds represent early, middle, and late eluters over the



range of the Consent Decree compounds and are not, themselves,



included on the list.  Construct Quality Control charts from



the data obtained according to directions in Reference 9.





     The sample matrix can affect the purging efficiencies



of individual compounds; therefore, each sample must be



dosed with the internal standards and analyzed in a manner



identical to the internal standards in blank water.  When



the results of the dosed sample analyses show a deviation



greater than two sigma, repeat the dosed sample analyses.



If the deviation is again greater than two sigma, dose



another aliquot of the same sample with the compounds  of



interest at approximately two times the measured values and



analyze.   Calculate the recovery for the individual compounds



using these data.*
*See Reporting of Data Section,  p.  11.

-------
                           -  10  -

     Calibration of the gas chromatography-mass spectrometry

(GC-MS) system - Evaluate the system performance each day

that it is to be used for the analysis of samples or blanks.

Inject a sample of 20 nanograms of decalfuorotriphenyl-

phosphine    and plot the mass spectrum.  The criteria in

Reference  2  must be met and all plots from the performance

evaluation, documented and retained as proof of valid

performance.                         "   ?


     Analyze the 20 yg/1 standard to demonstrate instrument

performance for these compounds.


     Qualitative and Quantitative Determination - The char-

acteristic masses or mass ranges listed in Table II of this

section are used for qualitative and quantitative determination

of volatile priority pollutants.  They are used to obtain an
                                     (e)
extracted ion current profile (EICP)    for each compound.

For very low concentrations, the same masses may be used for

selected ion monitoring  (SIM)    .  The primary ions to be used

to quantify each compound are also listed.  If the sample pro-

duces  an interference for the primary ion, use a secondary

ion to quantify-
 (d)  Available from PCR, Inc., Gainesville, FL.

 (e)  EICP is the reduction of mass spectrometric data
     acquired by continuous, repetitive measurement of
     spectra by plotting the change in relative abundance
     of one or several ions as a function of time.

 (f)  SIM is the use of a mass spectrometer as a substance
     selective detector by measuring the mass spectrometric
     response at one or several characteristic masses in
     real time.

-------
                                - 11 -



          Quantify samples by comparing  the  area of a single mass



     (see Table II) of the unknown  in  a  sample  to that of a stan-



     dard.  When positive responses are  observed,  prepare and



     analyze a reference standard so that the standard response



     closely approximates the sample response.   Calculate the con-



     centration in the sample as follows:



               (Area for unknown)

               (Area for standard)                _    ,,   - ,,„,,„--„,
                7t	r—rr~-	=— j.  j   j  i   /i \    U9/1 °f unknown
                Concentration or standard  (vig/1)     3




5.   Reporting of Data



          Report all results to two significant  figures or to the



     nearest 10 yg/1.  Report internal standard  data to two  signif-



     icant figures.



          As the analyses are completed, transfer GC-MS data  to



     magnetic tape .as described under reporting  of data in method


     for "Organics by Liquid-Liquid Extraction - Gas Chromatography."




          Report all quality control (QC) data along with  the



     analytical results for the samples.  In addition, forward



     all QC data to EMSL, Cincinnati.




6.   Direct Aqueous Injection Gas Chromatography



          As noted in the Scope, Acrolein and acrylonitrile should



     be analyzed by direct aqueous injection gas chromatography-



     mass spectrometry.   See references (3) ,  (4)  -, and  (5)  for



     these methods.   The  detection level for these methods is 0.1


     mg/1 and above.

-------
                           - 12 -
                           Table I
         Elution Order of Volatile Priority Pollutants
  (a)
Compound
chlorome thane"
dichlorodifluoromethane
bromomethane
vinyl chloride
chloroethane
methylene chloride
trichlorofluoromethane
1,1-dichloroethylene
bromochloromethane(IS)
1,1-dichloroethane
trans-1,2-dichloroethylene
chloroform
1,2-dichloroethane
1,1,1-trichloroethane
carbon tetrachloride
bromodichloromethane
bis-chloromethyl ether
1,2-dichloropropane
trans-1,3-dichloropropene
trichloroethylene
dibromochloromethane
cis-1,3-dichloropropene
1,1,2-trichloroethane
benzene
2-chloroethylvinyl ether
2-bromo-l-chloropropane(IS)
bromoform
1,1,2,2-tetrachloroethene
1,1,2,2-tetrachloroethane
                                  (b)
Purging
Efficiency
(percent)
Purging
Efficiency
Modified
Method
(percent)
0.152
0.172
0.181
0.186
0.204
0.292
0.372
0.380
0.457
0.469
0.493
0.557
0.600
0.672
0.684
0.750
0.760
0.818
0.847
0.867
0.931
0.913
0.913
0.937
0.992
1.000
1.115
1.262
1.281
91
0 100 (c)
85
101
90
76
96
97
88
89
92
95
98
94
87
92
0
92
90
89
87
85
88
no data
no data
92
71
88
58

-------
                           - 13 -
                    Table I  (cont'd)
Compound

1,4-dichlorobutane(IS)
toluene
chlorobenzene
ethylbenzene
acrolein
acrylonitrile
 RRT
                                 (b)
 1.312

 1.341

 1.489
 1.814

unknown

unknown
Purging
Efficiency
(percent)

   74
 no data
   89
 no data
   12
 no data
Purging
Efficiency
Modified
Method
(percent)
   74
     (e)
(a)   These data were obtained under the following conditions
     GC column - stainless steel, 8 ft. long x 0.1 in. I.D.
     packed with Carbopack C (60/80 mesh), coated with 0.2%
     Carbowax 1500;  preceeded by a 1 ft. long x 0.1 in. I.D.
     column packed with Chromosorb W coated with 3% Carbowax
     1500; carrier flow - 40 ml/min.; oven temperature -
     initial 60°C held for 3 min., programmed 8°C/min. to
     160°C and held until all compounds eluted.  The purge
     and trap system used was constructed by EPA.  Under
     optimized conditions, commercial systems will provide
     equivalent results.

(b)   Retention times relative to 2-bromo-l-chloropropane
     with an absolute retention time of 829 seconds.

(c)   No measurable recovery using standard purging and trap-
     ping conditions.  Under modified conditions, i.e.,
     purging at 10 ml/min. for 12 min., recovery is 100%.

(d)   Bis-chloromethyl ether has a very short half-life in
     water and is not likely to be detected in water.

(e)   Recovery 12% under standard purging conditions, i.e.,
     room temperature,  30% at 55°C,  and 74% at 95°C.

-------
                                - 14 -
                                  l'
                                Table II
              Characteristic Ions of Volatile Organics
Compound
chloromethane
dichlorodi fluoromethane

bromomethane
vinyl chloride
chloroethane
methylene chloride

trichlorofluoromethane
1,1-dichloroethylene
bromochloromethane(IS)

1,1-dichloroethane

trans-1,2-dichloroethylene
chloroform
1,2-dichloroethane

1,1,1-trichloroethane

carbon tetrachloride
bromodichloromethane

bis-chloromethyl ether
• 1,2-dichloropropane

trans-1,3-dichloropropene
trichloroethylene

dibromochloromethane

cis-1,3-dichloropropene
El Ions  (Relative
   intensity)
                 Ion used to
                  quantify
50(100) ; 52(33)
85(100) ; 87(33) ;
101(13); 103(9)
94(100) ; 96(94)
62(100) ; 64(33)
64(100) -, 66(33)
49(100>;51(33) ;
84(86}; 86(55)
101(100) ; 103(66)
61(100) ; 96(80) ; 98(53)
50
101
94
62
64
84
101
96
49(100); 130(88);
128(70); 51(33)
63(100); 65(33); 83(13);
85(8); 98(7); 100(4)
61(100)
83(100)
 96(90)
 85(66)
98(57)
62(100); 64(33);
98(23) ; 100(15)
98(100); 99(66);
117(17); 119(16)
117(100); 119(96); 121(30)
83(100)
127(13)
 85(66);
 129(17)
79(100); 81(33)
63(100)
112(4);
 65(33);
114(3)
75(100): 77(33)
95(100)
130(90)
 97(66);
 132(85)
129(100) ; 127(78)
208(13) ; 206(10)
75(100); 77(33)
128

 63
 96
 83

 98

 97
117

127
 79

112
 75

130

127
 75

-------
                                - 15 -
Compound
1f1,2-trichloroethane

benzene
2-chloroethylvinyl ether
2-bromo-l-chloropropane(IS)
bromoform

1,1,2,2-tetrachloroethene

1,1,2,2-tetrachloroethane

1,4-dichlorobutane(IS)
toluene
chlorobenzene
ethylbenzene
acrolein

acrylonitrile
Table  (cont'd)
 El Ions  (Relative
    intensity)	
 83(95); 85(60); 97(100);
 99(63); 132(9); 134(8)
 78(100)
 63(95); 65(32); 106(18)
 77(100);  79(33);156(5)
 171(50) ,-173(100) ; 175(50)
 250(4); 252(11);  254(11);
 256(4)
 129(64)
 164(78)
131(62);
166(100)
 83(100);  85(66);  131(7)
 133(7);  166(5);  168(6)
 55(100);  90(30);  92(10)
 91(100);  92(78)
 112(100);  114(33)
 91(100);  106(33)
 26(49) ;  27(100);
 55(64);  56(83)
 26(100);  51(32);
 52(75);  53(99)
                 Ion  used to
                 quantify
 97
 78
106
 77

173

164

168
 55
 92
112
106

 56

 53

-------
                                -  16  -



                Organics by Liquid-Liquid Extraction -

                         Gas Chromatography



1.   Scope

          This method is designed to determine those "unambiguous

     priority pollutants" associated with the Consent Decree, that

     are solvent extractable and amenable to gas Chromatography.

     These compounds are listed in Tables III to V of this section.

     Except for the pesticides, it is a gas chromatographic-mass

     spectrometric method intended for qualitative and semi-

     quantitative determination of these compounds during the

     survey phase fof the industrial effluent study.  Pesticides

     are initially determined by electron capture-gas Chromatography

     and, qualitatively, confirmed by mass spectrometry.


2.   Special Apparatus and Materials


          Separatory funnels - 2 and 4-liter with Teflon stopcock

          Continuous liquid-liquid extractors - any such apparatus

               designed for use with solvents heavier than water

               and having a capacity of 2 to 5-liters   .   Con-

               necting joints and stopcocks must be of Teflon or

               glass with no lubrication.


3.   Procedure


          Sample Preparation for GC-MS Survey - Blend the  com-

     posite sample to provide a homogeneous mixture including
     (a)  Available from Aldrich Chemical Co., Milwaukee, WI,
          Catalog No. Z10, 157-5.

-------
                           - 17 -


a representative portion  of the  suspended  solids  that are


present.  No specific method is  required but  a motor driven


mechanical stirrer with a propeller  type blade is suggested.


Stirring with metal devices is acceptable  for organic sampling.


     Transfer the sample  from the composite container through
                                                       j
a glass funnel into a 2-liter graduated cylinder  and measure

the volume.  Then transfer  to a  4-liter separatory funnel or


a continuous extractor as described  below.  Rinse  the cylinder


with several portions of  the  first volume  of  extracting sol-

vent.   Note: [Either separatory funnel or continuous ex-

traction is acceptable for  isolation of the organics.  Contin-

uous extraction must be used  when emulsions cannot be broken.

See discussion under Emulsions.]



Base-Neutral Extraction


     Separatory Funnel Extraction - Adjust the pH of  the sample

with 6 N NaOH to 11 or greater.  Use multirange pH paper for


the measurement.  Serially extract with 250 x 100 x 100 ml

portions of distilled-in-glass methylene chloride.   (About 40 ml

of the first 250 ml portion will dissolve in the sample and not

be recovered.)   Shake each extract for at least 2 min by the

clock.


     Dry and filter the solvent extract by passing it through

a short column of sodium sulfate.  Concentrate the solvent by

Kuderna-Danish (K-D)  evaporation (distillation).   The sodium

sulfate should be prewashed in the column with methylene

-------
                           - 18 -
chloride.  [Note:  Check sodium sulfate blank and/  if
necessary, heat in an oven at 500°C for 2 hours to  remove
interfering organics.]  After drying the extract, rinse the
sodium sulfate with solvent and add to the extract.
     Evaporate the extract to 5-10 ml in a 500 ml K-D apparatus
fitted with a 3-ball macro-Snyder column and a 10 ml calibrated
receiver tube.  Allow the K-D to cool to room temperature.
Remove the receiver, add fresh boiling chips, attach a two-
chamber micro-Snyder column and carefully evaporate to 1.0 ml
or when active distillation ceases.  Remove the micro-Snyder
column and carefully evaporate to 1.0 ml or when active dis-
tillation ceases.  Remove the micro-Snyder column and add the
internal standard:  10 yl of 2 vg/yl d,Q-anthracene  (per each
ml of extract).  Mix thoroughly-
     If it is to be overnight or longer before the  extract is
run by GC-MS, transfer it from the K-D ampul with a disposable
pipet to a solvent tight container.   The recommended container
is a standard 2 ml serum vial with a crimp cap lined with
Teflon coated rubber.  These are inert and methylene chloride <
can be held without evaporation loss for months if  caps are
unpierced.  When the extracts are not being used for analysis,
store them with unpierced caps in the dark and at refrigerator
or freezer temperatures.
     Acid  (Phenols) Extraction - Adjust the pH of the base-
neutral extracted water with 6 N HC1 to 2 or less.  Serially

-------
                           - 19 -




extract with 200 x 100 x 100 ml portions of distiiled-in-



glass methylene chloride.   (Note that only 200 ml is used



for the first extraction).  Proceed as described for the base-



neutral extract, including the addition of the internal



standard.




     Emulsions - The recovery of 85% of the added solvent



will constitute a working definition of a broken emulsion.



(You may correct the recovery of the first portion for water



solubility of methylene chloride.)   Any technique that meets



this criteria is acceptable.  Among techniques that have been



tried on these samples with fair success are:



     1.   Centrifugation of the emulsion layer after removel



          of any separated solvent.



     2.   Passage of the emulsion through a column plugged



          with a ball of methylene  chloride-wet glass wool.



          The solvent used to wet the wool and to wash it



          after the emulsion goes through must be measured



          and subtracted from the total volume to determine



          85% recovery.



     3.   Relative to labor,  solvent is cheap.  The  addition



          of excess solvent sometimes breaks weak emulsions.



          You must remember to use  excess solvent in the



          blanks also,



     4.   Let the emulsion  stand  for up to 24  hrs.



     5.   Draw off the  small  amount of free solvent  that sep-



          arates and  slowly drip  it back  in the top  of  i;he

-------
                           -  20  -



          separatory funnel and through the sample and



          emulsion.



     Other ideas include stirring with a glass rod, heating



on a steam bath, addition of concentrated sodium sulfate



solution, and sonication.  See  discussion in Appendix  I.




     Continuous Extraction - If you cannot achieve 85% solvent



recovery, start with a fresh aliquot of sample and extract by



continuous extraction.




     Adjust the pH of the sample as appropriate, pour into
              <.


the extractor, and extract for 24 hours.  When extracting a



2-liter sample, using the suggested equipment, two liters of



blank water must be added to provide proper solvent recycle.




     For operation, place 200-300 ml of solvent in the ex-



tractor before the sample is added and charge the distilling



flask with 500 ml of solvent.  At the end of the extraction



remove the solvent from the distilling flask only and evap-



orate and treat as described in the base-neutral extract



section.




     Blank Extraction:  It is not entirely certain that



2 liters of blank will always be available.  When it is,



proceed to process it as the corresponding sample was done.



Include any emulsion breaking steps that used glass wool,



excess solvent or additional chemicals.  If less than 2 liters



is available, measure the blank and bring it to volume with



distilled water.  On analysis make the necessary quantita-



tive corrections.

-------
                           - 21 -




     Pesticides:  These compounds are to be analyzed by



EC-GC using the EPA method  published in the Federal  Register,



Vol. 38, Number 125, Part II, pp. 17318-17323.   (Friday,



June 29, 1973).  One-liter  rather than 100 ml is  to  be  ex-



tracted.  The solvent amounts given in the method and other



parameters remain unchanged.  If pesticides are found by EC,



the extract is to be carefully evaporated  (clean  airstream)



to 0.5 ml and sent for GC-MS confirmation.




     The compounds to be analyzed by EC-GC are listed in



Table III.




     If the pesticide sample has been received in a  1-gal.



bottle, hand shake the bottle for 1 rain,  by the clock to evenly



suspend sediment.  Pour the sample into a 1-liter graduated



cylinder and measure the volume.  Then transfer the  sample



to a 2-liter separatory funnel and rinse the cylinder with



the first volume of extracting solvent.   Use additional small



volumes of solvent if necessary to transfer all of the sample.



Proceed with the extraction using the solvents and amounts



prescribed in the published method.




     If the sample is to be taken from the original composite



bottle, homogeneously mix as described earlier and transfer



a 1-liter aliquot to a graduated cylinder,  then transfer



to the separatory funnel with the aid of  a glass funnel 'and



rinse the cylinder as above.

-------
                           - 22 -


     if intractable emulsions  are encountered that cannot

be broken as described in the  GC-MS survey section/ then a

fresh 1-liter sample should be processed in a continuous

extractor using methylene chloride as the solvent as des-

cribed earlier.  The methylene chloride will have to be

evaporated to a small volume and exchanged into hexane for

clean-up or EC-GC analysis.  To do this, evaporate the methy-

lene chloride to 6 to 8 ml, cool, add 20 ml of hexane and

a fresh boiling stone and re-evaporate to the desired analy-

tical volume ((5 ml or less) .


     Final storage and transport of sample extracts:  After

analysis, the extracts of the  base-neutrals, acids, blanks

and pesticides are to be sent  to ERL, Athens, GA  30601,

ATTN:  Dr. Walter Shackelford.


     Each extract is to be washed out of its container into

a 10 ml glass ampul and brought to 5 ml ± 1 ml.  Methylene

chloride is the solvent for the base-neutrals and acids,
                                 •
hexane for pesticides.  The ampuls are to be sealed in a

rounded-off, fire polished manner, i.e., no thin sharp peaks

of glass that are easily broken on handling and shipping.

After sealing the ampuls, put  an indelible mark at the

solvent level.  Securely attach a label or tag that gives:

          Type of fraction  (base-neutral, etc.)

          Industrial category

          Name (of plant, city and state)

-------
                           - 23 -




          Specific  source  or  stage  of  treatment



          Date sampled originally



          Date sealed                                  .
                                                       i


          Name of contractor  and analytical  laboratory




     Wrap the ampuls in packing material to  prevent breakage



and mail or ship them postpaid at ambient temperature.  When



the samples are safely in  ampuls/ the remainder of the com-



posite sample may be discarded.





GC-MS Analysis





     Compounds to be analyzed by GC-MS alone fall into two



categories—those in the base-neutral extract  (Table iv) and



those in the acid extract  (Table  V ).  Pesticides (Table III



that were tentatively identified in the pesticide analysis



will be confirmed by GC-MS.




     The base-neutral extractables may be separated and eluted



into the MS under the following chromatographic conditions:




          Column - 6 foot,  2.0 mm inside diameter,- glass



          Packing - 1%  SP2250 on 100/120 mesh Supel.copbrt



          Program - hold 4  minutes @ 50°, program 50°-260°



               § 8°/min.,  hold 20 minutes @260°



          Injector - 275°



          Separator - 275°



          Carrier gas -  He  @ 30 ml/min



          Injection size -  >2  ul

-------
                           - 24 -



     Table IV lists the 49 base-neutral extractable compounds


in order of relative retention times  (compared to hexachloro-


benzene) for the above GC conditions.  Detection limits were


determined by MS response.  The seven compounds without re-


tention times or limits of detection were not available for


this report.  It is not recommended that 2,3,7,8-tetrachloro-


dibenzo-p-dioxin (TCDD) be acquired due to its extreme tox-


icity.  Based on their similarity to compounds that were avail-


able all seven are expected to be chromatographable using


these standard conditions.  In addition the characteristic


masses recommended for MS identification are listed in Table IV.



     The limits of detection given in Tables ill and IV refer


to the quantity necessary to inject to get confirmation by


the MS methods described below.



     At the beginning of each GC-MS run of a base-neutral


extract, the operator should demonstrate the ability to chro-


matograph benzidine at the 40 ng level.  Only after this is


accomplished should the run be started.  If benzidine can be


chromatographed, the other nitrogen-containing compounds of


Table IV can be chromatographed as well.



     If desired, capillary or SCOT columns may be used instead


of the packed column of SP-2250.  Coatings of OV-17 or SP-2250


may be used.  The elution order of OV-17 and SP-2250 are very


similar.  Some specific data for OV-17 is given in Table VII.
                                 •

The performance criteria for benzidine must still be met

-------
                           - 25 -




and in addition, the  system must be  shown  to  elute  the late



running polynuclear aromatic compounds.




     The acid extractables may be chromatographed as  follows:




          Column, 6 foot, 2.0 mm inside diameter, glass



          Packing - Tenax GC, 60/80  mesh



          Program - 180° - 300° @ 8°/min



          Injector -  290°



          Separator - 290°



          Carrier Gas - He @ 30 ml/min



          Injection size - >2 ul




     Table  V  lists  the 11 acid extractables in order of



relative retention times (compared to 2-nitrophenol).  Chroma-



tography of nitrophenols is poor.  The limits of detection



given refer to the amounts required  to get MS confirmation



by the methods described below.  See Appendix I.




     Before an acid extract is run on the GC-MS the operator



should demonstrate the ability to detect 100 ng of>penta-



chlorophenol.



     Mass Spectrometry should be conducted with a system



utilizing a jet separator for the GC effluent since membrane



separators lose sensitivity for light molecules and glass



frit separators inhibit the elution of polynuclear aromatics.



A computer system should be interfaced to the mass spectro-



meter to allow acquisition of continuous mass scans  for the



duration  of the chromatographic program.   The computer system

-------
                           -  26  -
should also be equipped with mass storage devices for
saving all data from GC-MS runs.  There should be computer
software available to allow searching any GC-MS run for
specific ions and plotting the  intensity of the ions with
respect to time or scan number.  The ability to integrate
the area under any specific ion plot peak is essential for
quantification.
     To indicate the presence of a compound by GC-MS, three
conditions must be met.  First, the characteristic ions for
the compound ,(Tables m-v) must be found to maximize in the
same spectrum.  Second, the time at which the peak occurs
must be within a window of ± 1 minute for the retention time
of this compound.  Finally, the ratios of the three peak
heights must agree with the relative intensities given in
Tables III-V within ± 20%.
     An example of identifying  a component is as follows:
     It is known that hexachlorobenzene elutes from the SP2250
column at 19.4 minutes.  Hexachlorobenzene has characteristic
mass ions at 284(100%), 142(30%), and 249(24%).  The computer
is asked to display a plot of the intensities of these ions
versus time  (or MS scan number) and the window from 18.4-20.4
minutes is examined for the simultaneous peaking of the in-
tensities of these ions.  If all three ions are present, the
ratio of the peak heights is checked to verify that it is
100:30:24 ± 20%.  If the three tests are successful, hexachloro-
benzene has been identified in the sample.

-------
                           - 27 -




     Table III lists  the  18  pesticides  and  PCB's  that will be



confirmed by  GC-MS  using the  SP2250 column.   Chlordane,



toxaphene and the PCB's have  retention  ranges rather than



specific times due  to their being multicomponent mixtures.



It is suggested that the first 14 materials be  confirmed



exactly as the other base-neutral compounds.




     The last four  materials  require  special  treatments.  Chicr-



dane is expected to produce two main  peaks within the retention



range given in which all three masses listed  will maximize.



Toxaphene will produce several  (5-15) peaks in which the  masses



given will maximize within the retention  time range.  For the



PCB's each mass given corresponds to  the molecular  ion of PCB



isomers, e.g., 294  is tetrachlorobiphenyl.  A specific mass plot



will show multiple peaks for each of  these ions within the re-



tention time  listed, but in general they will not maximize in



the same TIC peak.  For these four materials  in particular it



is necessarv to also run a standard.  Because GC-MS is only



being used for confirmation—and at its limit of detection—all



quantification will be done by EC-GC  for the pesticides.  The



methods for these four are not final and feedback from the



field to Dr.  Shackelford is welcome.




     When a compound has been identified/  the quantification



of that compound will be based on the integrated area from



the specific  ion plot of the first listed characteristic ion



in Tables IV  and  V.   Quantification will be done by the



internal standard method using deuterated anthracene.  Response

-------
                           - 28 -

factors, therefore, must be calculated to compare the MS

response for known quantities of each priority pollutant with

that of the internal standard.  The response ratio  (R) may

be calculated as:

              Ac   Ca
          R " Aa x Cc

where Ac is the integrated area of the characteristic ion from

the specific ion plot for a known concentration, Cc.  Aa and

Ca are the corresponding values for deuterated anthracene.

The relative response ratio for the priority pollutants

should be known for at least two concentration values—40 ng

to approximate 10 ppb and 400 ng to approximate the 100 ppb

level.  Those compounds that do not respond at either of these

levels may be run at concentrations appropriate to their res-

ponse.  For guidance in MS limits of detection refer to the

values given in Tables III-V.

     The concentration of a compound in the extract may now

be calculated using:

          _ _ Ac x Ca
          C   Aa x R

where C is the concentration of a component, Ac is the inte-

grated area of the characteristic ion from the specific ion

plot, R is the response ratio for this component, Aa is the

integrated area of the characteristic ion in the specific

ion plot for deuterated anthracene, and Ca is the concentration

of deuterated anthracene in the injected extract.

-------
                                -  29  -
          In samples that contain an inordinate number of inter-
     ferences the chemical ionization (CI) mass spectrum may make
     identification easier.  In Tables iv and  v  characteristic
     CI ions for most compounds are given.  The use of chemical
     ionization MS to support El is encouraged but not required.

5.   Quality Assurance

          GC-MS system performance evaluation is required each day
     the system is used for samples or reagent blanks.  A sample
     of 20 ng of decafluorbtriphenylphosphine    is injected into
     the system and the mass spectrum is acquired and plotted.
     Criteria established in Reference 2 must be met.  The analyst
     must also demonstrate that the analytical conditions employed
     result in sharp total ion current peaks  for 40 ng of benzidine
     on the SP2250 column when this column is used and 100 ng of
     pentachlorophenol on the Tenax GC column when it is used with
     the MS as a detector.  All plots from the performance evalu-
     ation must be retained as proof  of  valid performance.
          As performance evaluation samples become available from
     EMSL-Cincinnati,  they are to  be  analyzed by solvent extraction
     once each 20 working days and the results reported with other
     analytical data.
          The 1% SP2250 and Tenax  GC  column packings  are available
     by request to EPA contractors from  Dr. Walter Shackelford,  EPA,
     Athens,  GA.
     (b)   Available  from PCR, Gainesville, FL

-------
                                - 30 -

          Standards for the priority pollutants may be obtained

     from the sources listed  in  Appendix 1^.   Those compounds

     marked with an asterisk  have  not yet been received by the

     Athens laboratory.

          In order to minimize unnecessary GC-MS analysis of blanks,

     the extract may be run on a FID-GC equipped with appropriate

     SP2250 and Tenax GC columns.   If no peaks are seen of intensi-

     ties equal to or greater than the deuterated anthracene internal

     standard, then it is not necessary to do a GC-MS analysis.  If

     such peaks are seen, then the blank must be sent for full

     priority pollutant analysis.

          The contractor will look for all priority pollutants to

     the limit of 10 yg/1 except in those cases listed in Tables iv-V

     in which limits of detection  are too high for analysis at this

     level.
                                                   ;

6.   Reporting of Data


          All concentrations  should be reported in ranges—10 ppb,

     100 ppb, and greater than 100 ppb.  Report concentrations for

     pesticides as prescribed in the Federal  Register Method.  The

     relative response ratios from MS analysis should be included  •

     when reporting data.                                           )

          All GC-MS data is to be  saved on 9-track magnetic tape

     and sent to the Athens Environmental Research Laboratory for

     storage and later evaluation.  The tape  format is:

        *Those labs which are under contract to perform GC-MS analyzes
        for EPA, may obtain a set  of standards from Mr. William Telliard,
        Chief, Energy'and Mining Branch, (202) 426-2726.

-------
                           - 31 -




          Type - 9 track,  300  BPI, 2400 foot reels



          Record length -  80



          Block Size - <4QQQ  (specify)



          Code - EBCDIC




     An acceptable data format would have the first two records



containing the sample identification.  Subsequent records con-



tain eight mass-intensity pairs, each of which is 10 characters



long.  Each mass and each intensity is 5 characters long and



left justified.  At the end of each spectrum in a sample run,



the last mass-intensity pair is blank to denote the end of the



spectrum.  When all data for the run is on the tape, an  end-



of-file mark should be written.  The next sample run can then



be entered.  One example is:



     2 Records:Sample 1 identification



     N Records:Spectrum 1 of sample,  last mass-intensity



               pair is blank to denote end of spectrum



     M Records:Spectrum 2 of sample,  last mass-intensity



               pair is blank to denote end of spectrum
     L Records:Spectrum N of sample,  last mass-intensity



               pair is blank to denote end of spectrum,



     END OF FILE



     2 Records:Sample 2 identification



     etc.

-------
                          - 32 -

     Other data formats are possible, but any format that is

used must be accompanied by a full explanation of all record

formats.


     All magnetic tapes, documentation and a table of MS res-

ponse ratios should be sent to:

          Dr. W. M. Shackelford
          Athens Environmental Research Laboratory
          College Station Road
          Athens, GA  30601

-------
                                Table III.
                     - 33 -
                     Pesticides
  Compound Name
6-endosulfan
a-BHC
Y-BHC
S-BHC•
aldrin
heptachlor
heptachlor epoxide
a-endosulfan
dieldrin
4,4'-DDE
4,4'-DDD
4,4'-DDT
endrin
endosulfan sulfate

6-BHC
chlordane
toxaphane
PCB-1242
PCB-1254
        RRT-1-
(hexachlorobenzene)
0
1
1
1
1
1
1
. 1
1
1
1
1
1
1
1
1
0
1
.51
.02
.09
.12
.14
.15
.23
.24
.28
.30
.33
.38
.41
.41
.14-1.37
.22-1.47
.93-1.24
.18-1.41
Detection Limit
     (ng)	
                               40
                               40
                               40
                               40
                               40
                               40
                               4u
                               40
                               40
                               40
                               40
                               40
                               40
                               20
 Characteristic
El ions (Rel. Int.)
                    201(100)
                    183(100)
                    183(100)
                    181(100)
                    66(100),
                    100(100)
                    355(100)
                    201(100)
                    79(100),
                    246(100)
                    235(100)
                    235(100)
                    81(100),
                    272(100)
        , 283(48)
        , 109(86)
        , 109(86)
        , 183(93)
         220(11),
        , 272(60)
        , 353(79)
        , 283(48)
         263(28) ,
        , 248(64)
        , 237(76)
        , 237(72)
         82(61),
        , 387(75)
,  278(30)
,  181(91)
,  181(91)
,  109(62)
 263(73)
,  274(46)
,  351(60)
,  278(30)
 279(22)
,  176(65)
,  165(93)
,  165(59)
263(70)
,  422(25)
                                             183(100), 109(86), 181(90)
                                             373(19),   375(17),   377(10)
                                             (231, 233, 235)*
                                             (224, 260, 294)*
                                             (294, 330, 362)*
*  These ions are listed without relative intensities since the mixtures they represent
   defy characterization by three masses.
**  These three ions are characteristic for the a and y forms of chlordane.  No stock
   should be set in these three for other isomers.

   1% SP-2250 on 100/120 mesh Supelcoport in a 6' x 2 mm" id glass column; He @ 30 ml/min;
   Program:  50  for 4 min, then 8 /min to 260  and hold for 15 min.

-------
                                                     34 -
                                 Table.IV.   Base-neutral Extractables
   Compound Name
(hexachloro-
   benzene)
 1, 3-dichlorobenzene
 1,4-dichlorobenzene
 foexachloroethane
 1,2-dichlorobenzene
 bis (2-chloroisopropyl)
   ether
 hexachlorobutadiene
 1,2,4-trichlorobenzene
 naphthalene
 bis(2-chloroethyl)ether
 hexachlorocyclopentadiene
 nitrobenzene
 bis(2-chloroethoxy)methane
 2-chloron aphthalene
 acenaphthylene
 acenaphthene
 isophorone
 fJuorene
 2,6-dinitrotoluene
 1,2-diphenylhydrazine
 2,4-dinitrotoluene
 N-ni trosod iphenylamine
 hexachlorobenzene
 4-bromophenyl phenyl ether
 phenanthrene
 anthracene
dimethylphthalate
diethylphthalate
 fluoranthene
pyrene  It
di-n-butylphthalate
benzidine
 butyl benzylphthalate
     0.35
     0.36
     0.38
     0.39

     0.47
     0.55
     0.55
     0.57
     0.61
     0.64
     0.64
     0.68
     0.76
     0.83
     0.86
     0.87
     0.91
    -0.93
     0.96
     0.98
     0.99
    1,
    1,
    1,
    1,
    1,
  00
  01
  09
  09
  10
1.15
1.23
1.30
1-.31
1.38
1.46
Limit of
Detection
  (ng)

   40
   40
   40
   40

   40
   40
   40
   40
   40
   40
   40
   40
   40
   40
   40
   40
   40
   40
   40*
   40
   40*
   40
   40
   40
   40
   40
   40
   40
   40
   40
  40*
  .40
                                Characteristic
                               El ions (Rel.  Int.)
                           146(100),
                           146(100),
                           117(100),
                           146(100),
           148(64),
           148(64),
           199(61),
           148(64),
          113(12)
          113(11)
          201(99)
          113(11)
 45(100)',
 225(100,
 74(100),
 128(100)
 93(100),
 237(100)
 77(100),
 93(100),
 162(100)
 152(100)
 154(100)
 82(100),
 166(100)
 165(100)
 77(100),
 165(100)
 169(100)
 284(100)
 248(100)
 178(100)
 178(100)
 163(100)
 149(100)
 202(100)
 202(100)
 149(100)
184(100).
149(100)
                                    77CU9),  79(12)
                                    223(63),  227(65)
                                    109(80),  145(52)
                                   ,  127(10),  129(11)
                                    63(99),  95(31)
                                   ,  235(63),  272(12)
                                    123(50),  65(15)
                                    95(32),  123(21)
                                   ,  164(32),  127(31)
                                   ,  153(16),  151(17)
                                   ,  153(95),  152(53)
                                    95(14),  138(18)
                                   ,  165(80),  167(14)
                                   ,  63(72),  121(23)
                                    93(58),  105(28)
                                   ,  63(72),  121(23)
168(71)
142(30),
250(99),
179(16),
179(16),
164(10),
178(25),
101(23),
101(26),
150(27),
167(50)
249(24)
141(45)
176(15)
176(15)
194(11)
150(10)
100(14)
100(17)
104(10)
                                      ,92(24),
                                      ,  91)50)
      185(13)
     CI  ions
    (Methane)

 146,  148,  150
 146,  148,  150
 199,  201,  203
 146,  148,  150

 77,  135, 137
 223,  225,  227
 181,  183,  209
 129,  157,  169
 63, 107, 109
 235,  237,  239
 124,  152,  164
 65, 107, 137
 163,  191,  203
 152,  153,  181
 154,  155,  183
 139,  167,  178
 166,  167,  195
 183,  211,  223
 185,  213,  225
 183,  211,  223
 169,  170,  198
 284,  286,  283
 249,  251,  277
 178,  179,  207
 178,  179,  207
 151,  163,  164
 177,  223,  251
 203,  231,  243
 203,  231,  243
 149, 205,  279
185, 213, 225
149, 299, 327

-------
                                                  - 35 -
                              Table -'IV.  Base-neutral Extractables  (Cont'd.)
  Compound Name
(hexachloro-
   benzene)
chrysene
bis (2-ethylhexyl)phthalate
benzo(a)anthracene
benzo (b)fluoranthene
benzo(k)fluoranthene
benzo(a)pyrene
indeno(1,2,3-cd)pyrene
dibenzo(a,h)anthracene
benzo(g h i)perylene
       46
       50
       54
       66
       66
       73
       07
     2.12
     2.18
N-nitrosodimethylamine
N-nitrosodi-n-propylamine
4-chloro-phenyl phenyl ether
endrin aldehyde
3,3' -dichlorobenzidine
2,3,7,8-tetrachlorodibenzo-
  p-dioxin
bis (chioromethyl)ether
deuterated anthracene (dlO)  1.09
Limit of
Detection
  (ng)

   40
   40
   40
   40
   40
   40
  100
  100
  100
  Characteristic
El ions (Rel. Int.
                    40
228 (100)
149(100)
228,(100)
252(100)
252(100)
252(100)
276 (100)
278(100)
276(100)
42 (100) ,
130(22) ,
204(100)
252(100)
322(100)
45(100),
188(100)
, 229(19)
, 167(31)
, 229(19)
, 253(23)
, 253(23)
, 253(23)
, 138(28)
, 139(24)
, 138(37)
74(88), t
42(64), ]
, 206(34)
, 254(66)
, 320(90)
49(14) ,
, 94(19),
226(23)
279(26)
226 (19)
125(15)
125(16)
125(21)
277(27)
279(24)
277(25)
14(21)
L01(12)
, 141(29)
, 126(16)
, 59(95)
51(5)
80(18)
    CI ions
   (Methane)

228,  229, 257
149 '
228,  229, 257
252,  253, 281
252,  253, 281
252,  253, 281
276,  277( 305
278,  279, 307
276,  277, 305
                                             189,  217
   1% SP-2250 on 100/120 mesh Supelcoport in a 61 x  2 mm id glass  column;  He  @  30 ml/min;
   Program:  50° for 4 min, then 8 /min to 260  and  hold for  15 min.

   Conditioning of column with base  is  required.                    - -	

-------
                                                  - 36  -
                                       Table  V.    Acid Extractables
  Compound Name
      RRT
(2-nitrophenol)
2-chlorophenol   ,   .         ,0..63
phenol                        0.66
2,4-dichlorophenol            0.96
2-nitrophenol                 1.00
p-chloro-m-cresol             1.05
2,4,6-trichlorophenol      '   1.14
2,4-dimethylphenol            1.32
2,4-dinitrophenol             1.34
4,6-dinitro-o-cresol          1.42
4-nitrophenol                 1.43
pentachlorophenol             1.64
deuterated anthracene  (dlO)   1.68
Limit of
Detection
  (ng)

  100
  100
  100
  100
  100
  100
  100
  2 yg
  2 vi g
  100
  100
   40
Characteristic
El ions (Rel. Int.)
128(100), 64(54), 130(31).
94(100), 65(17), 66(19)
162(100), 164(58), 98(61)
139(100), 65(35) , 109(8)
142(100) , 107(80), 144(32)
196(100), 198(92), 200(26)
122(100), 107(90), 121(55)
184(100), 63(59) , 154(53)
198(100), 182(35), 77(28)
65(100), 139(45) , 109(72)
266(100), 264(62), 268(63)
188(100), 94(19), 80(18)
CI ions
(Methane)
129, 131,^157
95, 123, 135
163, 165, 167
140, 168, 122
143, 171, 183
197, 199, 201
123, 151, 163
185, 213, 225
199, 227, 239
140, 168, 122
267, 265, 269
189, 217
   Column:  6' glass, 2  nun i.d.
            Tenax GC - 60/80 mesh
            180° - 300° 6 8°/min.
            He <§ 30 ml/ruin

-------
                             - 37 -
     Table VI.  ELUTION ORDER OF MOST OF THE SEMIVOLATILE
               PRIORITY POLLUTANTS ON 1% SP2250a
	Compound	RRT  'C
1,3-dichlorobenzene                             0.35
2-chlorophenol                                  0.35e
1,4-dichlorobenzene                             0.36
hexachloroethane                                0.38
1,2-dichlorobenzene                             0.39
bis(2-chloroisopropyl)ether •                    0.47
3-endosulfan                                    0.51
2,4-dimethyl, phenol                             0.52e
2-nitrophenol                                   0.53e
2,4-dichlorophenol                              0.53e
hexachlorobutadiene                             0.55
1,2,4-trichlorobenzene                          0.55
naphthalene                                     0.57
bis(2-chloroethyl)ether                         0.61
hexachlorocyclopentadiene                       0.64
nitrqbenzene                                    0.64
phenol                                          0.67
bis(2-chloroethoxy)methane                      0.68
2,4,6-trichlorophenol                           0.71e
p-chloro-m-cresol                               0.73
2-chloronaphthalene                             0.76
acenaphthylene                                  0.83
acenaphthene                                    0.86
isophorone                                      0.87
fluorene                                        0.91

-------
                             - 38 -
     Table VI.   ELUTIOM ORDER OF MOST OF THE SEMtVOLATlLE
         PRIORITY POLLUTANTS ON 1% SP22503  (Continued)
                                                   b c
	Compound   	^	           RRT ' _
2,6-dinitrotoluene                              0.93
1,2-diphenylhydrazine                           0.96
2,4-dinitrotoluene                              0.98
N-nitrosodiphen,ylamine                          0.99
hexachlorobenzene                               1.00
4-brorcophenyl phenyl ether                      1.01
a-BHC                                           1.02
                                                    f
Y-BHC                                      "     1.09
phenanthrene                                    1.09
anthracene     ,                                 1.09
dimethyl phthalate      '                        1.10
                                                    f
pentachlorophenol                               1.11
S-BHC                                           1.12
aldrin                                          1.14
diethyl phthalate                               1.15
heptachlor                                      1.15
heptachlor epoxide                              1.23
fluoranthene                                    1*23
ct-endosulfan                                    1.24f
dieldrin                                        1428;
4,4'-DDE                             •           1.30
pyrena                               '           1*30
di-n~butyl phthalate                            1.31
4,4'-DDD  (p,p'-TDE)                             1.33
4,4'—DDT                                       1.38d
endosulfan sulfate                              1.41
endrin                             .             1.41
benzidine                                     -  1.38
butyl benzyl phthalate                          1.46
chrysene                                        1.46

-------
                             - 39  -
     Table VI.  ELUTION ORDER OF MOST  OF THE SEMI VOLATILE
         PRIORITY POLLUTANTS  ON  1%  SP22503 (Continued)
            Coirtpound
•KRTbj
bis(2-ethylhexyl)phthalate                       1.50
benzo(a)anthracene                               1.54
benzo(b)fluoranthene                             1.66
benzo(k)fluoranthene                             1.66
benzo(a)pyrene                                   1.73
indeno(l,2,3-cd)pyrene                           2.07
dibenzo(a,h)anthracene                           2.12
benz o(ghi)pe rylene                               2.12
a 1% SP-2250 on 100/120 mesh Supelcoport in a 6f x 2irtm id
  glass column; He @ SOml/min; Program:  50° for 4 min,
  then 8°/min to 260° and hold for 15 min.

  Relative to hexachlorobenzene at 19.4 min.
  40ng gives 5-90% response on FID unless otherwise noted.

  200ng required to obtain 5-90% response on FID.

  2 11 g required.

  40 yg required.

-------
                            - 40 -

                        Table 'VI
                        (continued)


Standards not available;  as of 2/8/77
     N-nitrosodi-n-propylamine
     4-chlorophenyl phenyl ether
     TCDD
     endrin aldehyde
     N-nitrosodimethylaraine
     3,3' -dichlorobenzidine
     bis (chlorome thy 10 ether (unstable in water)

Standards that would not chromatograph;
     4 ,6-dinitro-o-cresol
     4 -ni tropheno 1
     2 , 4-dinitrophenol

Standards yielding a range of  eaks;
     PCB-1242              0.93-1.24
     PCB-1254              1.18-1.41
     toxaphene-             1 . 22-1 . 4 7
     chlordane             1.14-1.37

-------
                           - 41 -
          -  .._ Table VII.   Order of  Elution  for
                           OV-17 SCOT Column

                                                                 2
     Compound                                     Spectrum Number

1,3-dichiorobenzene                                    134
1,4-dichlorobenzene                                    137
2-chlorophenol                                         141
1,2-dichlorobenzene                               '     153
bis(2-chloroethy1)ether                                163
phenol                                                 165
bis (2-chloroisopropyl)ether                            173
hexachloroethane                                       178
nitrobenzene                            '               194
2-nitrophenol                                          219
1,2,4-trichlorobenzene                                 234
2,4-dimethylphenol                                     240
naphthalene                                            240
2,4-dichlorophenol                                     244
hexachlorobutadiene                                    262
isophorone                                             272
p-chloro-m-cresol                                      317
hexachlorocyclopentadiene                              325
2,4,6-trichlorophenol                                  332
chloronaphthalene                                      339
2,4-dinitrotoluene                                     372
acenaplithylene                                         374
acenaphthene                                           390
dimethylphthalate                                      397
fluorene                                               434
diethylphthalate                                       447
N-nitrosodiphenylamine                                 447
2,6-dinitrotoluene                      ^              454
a-BHC                                   '              476
4-bromophenyl phenyl ether                             478
T-BHC                                                  487
hexachlorobenzene                                      490
8-BHC                              .                    506
phenanthrene                                           518
anthracene                                             518
di-n-butylphthalate                                    583
aldrin                                                 592
fluoranthene              '                             617
pyrene                                                 634
DDE                                                    659
ODD                                            •        664
endrin                                                 688
dieldrin                                               688
DDT           .                                         713
butyl benzyl phthalate                                 713
benzo(a)anthracene                                     748
chrysene                                              748

-------
                           - 42 -

                    Table VII.  • continued
                                                            2
    Compound                                 Spectrum Number


bis(2-ethylhexyl)phthalate                            804
benzo(a)pyrene                                        906

banzo(b)fluoranthene                                  970
benzo(k)fluoranthene                                  970
   33 meter glass OV-17 SCOT column,
   Program:  60  -260  @6 /minute


2
   Number of 2.5 second scans up to point of elution.

-------
                               - 43 -
                              Metals





1.   Sample Preparation




          With the exception of mercury, the metals to be deter-



     mined may be divided into two groups as follow:



          a)    those metals which are to be first analyzed by



               flame atomic absorption (AA), and, if not detected,



               then analyzed by flameless AA—Be, Cd, Cr, Cu, Ni,



               Pb and Zn,



          b)    those metals which are to be analyzed by flameless



               AA only—Ag, As, Sb,  Se,  and Tl.




          For flame AA analysis the  sample should be prepared using



     the procedure as given in "Methods  for Chemical Analyses of



     Water and Wastes (1974)", 4.1.4, page 83 (Reference 7).




          With the exception of antimony and beryllium, samples to



     be analyzed by flameless AA should  be prepared as an industrial



     effluent as described in "Atomic Absorption Newsletter," 14,



     page 111 (1975)  (Reference 8).   Note:  Nickel nitrate should



     be added only to those aliquots on  which the analysis of



     selenium and arsenic are to be  accomplished.  The sample prep-



     aration  procedure for antimony  and  beryllium analysis by flame-



     less AA  is the same procedure used  for flame AA.




          The sample preparation procedure to be used for mercury



     analysis is that given in "Methods  for Chemical Analysis of



     Water and Wastes (1974)",  8.1,  page  124  (Reference 7).

-------
                                -  44  -

2.   Apparatus

          All samples are to be analyzed using an atomic absorption

     spectrophotometer equipped with simultaneous background

     capability.  For arsenic, cadmium, antimony, selenium, thallium,

     and zinc, either electrodeless discharge lamps or high intensity

     hollow cathode lamps may be utilized.  A heated graphite atom-

     izer is to be used for all flameless AA work.  A strip chart

     recorder must be used as part of the readout system to detect

     and avoid the inclusion of extraneous data.


3.   Procedure


          a)   Flame AA - The procedures to be used are those

               described in "Methods for Chemical Analysis of

               Water and Wastes (1974)"(Reference 7) as referenced

               in Table I below.  Instructions as to when flame-

               less AA is to be used are also included.  For

               those defined in the recommended procedures, the

               instrument manufacturers recommendations are to
                                                                  i
               be followed.  Background correction is to be used

               on all analyses.

-------
                           -  45  -

                         Table VIII

           Methods for Chemical
           Analysis of Water and
Element       Wastes, 1974*	           Comments

  Be             p. 99          Analyze by flameless AA if
                                cone. <20 yg/1

  Cd             p. 101         Analyze by flameless AA if
                                cone. <20 yg/1

  Cr             p. 105         Use nitrous oxide-acetylene
                                flame for all analyses—analyze
                                by flameless AA if cone. <200 yg/1

  Cu             p. 108         Analyze by flameless AA if
                                cone. <50 yg/1

  Ni             p. 141         Analyze by flameless AA if
                                cone. <100 yg/1

  Pb             p. 112         Analyze by flameless AA if
                                cone. <300 yg/1

  Zn             p. 155         Analyze by flameless AA if
                                cone. <20 yg/1
*In those instances where more vigorous digestion for sample
 preparation is desired (or necessary)  the procedure on page 82
 (4.1.3)  should be followed.

     b)    Standard solutions to be used for the flameless

          work should also be prepared as described in

          "Methods for Chemical Analysis of Water and Wastes

          (1974)" (Reference 7).  The working standards should

          be diluted to contain the same acid concentration as

          the prepared samples.  The instrumental settings

          and conditions recommended by the manufacturers are

          to be considered the procedural guidelines.   In

          addition,  the following requirements should also be

          incorporated into the procedures:

-------
                 -  46  -

1)   Argon should be used as the purge gas in
                                                   ,
     all analyses.

2)   Background correction and method of standard

     addition must be used on all analyses.

3}   A blank maximum temperature atomization, without

     gas interrupt, should be accomplished before

     each analytical determination.

4)   The graphite tube or cuvette should be replaced

     as suggested by the instrument manufacturer or

     when contamination or lack of precision indicates

     that replacement is necessary.

5)   All disposable pipet tips should be cleaned

     before use by soaking overnight in 5% redistilled

     nitric acid, rinsed with tap and deionized

     water, and dried.

6)   The accuracy of the temperature indicator on the

     heated graphite atomizer should be verified

     before beginning any analytical work.  This

     should be done by plotting charring temperature

     for a standard solution of a compound where the

     volatilization temperature is known.  The com-

     pound used should have a volatilization temper-

     ature between 800 and 1200°C.

7)   To insure that there is no loss from the acid ,
                                                   i

     matrix prior to atomization, the optimum charring

     temperature for each metal should be established

     in the same manner (i.e., by plotting charring

     temperature versus atomization signal of standard

     solution of each metal).

-------
                               - 47 -
          For the determination of selenium the procedure given
     for industrial effluents ("Atomic Absorption Newsletter,"
     Vol. 14, page 109 [1975])  (Reference 8) should be followed.
     Arsenic should be determined in the same manner (using the
     nickel nitrate matrix)  with an optimum charring temperature
     of approximately 1300°C.
          The analysis of zinc by flameless AA is difficult because
     of environmental contamination.  The analyst must take pre-
     caution to provide a clean work area to minimize this problem.
          c)    Mercury analyses - The cold vapor technique as
               described in  "Methods for Chemical Analysis of Water
               and Wastes, (1974)", page 118 (Reference 7) is to
               be followed.
4-.    Quality Assurance
          a)    To verify that the instrument is operating correctly
               within the expected performance limits,  an appropriate
               standard should  be included between every ten samples.
          b)    Spiked aliguots  shall be analyzed with a frequency
               of 15% of the  sample load for each metal determined
               by flame AA.   If the recovery is not within ±10% of
               the expected value the  sample should be  analyzed by
               method of standard addition.   (The spike should be
               added to the aliquot prior to sample preparation.)
               The amount added should  increase the absorbance by
               not less than  0.01 units where  the absorbance in the
               unspiked aliquot was less than  0.1,  and  not more than
               0.1 when the absorbance  in the  unspiked  aliquot was
               0.1 or greater.

-------
                               - 48 -




               c)    For mercury,  the spike added should be an



                    amount equal  to five times the detection level.



               d)    Reagent blanks shall be run for each metal



                    being determined with the sample values being



                    corrected accordingly.



               e)    When using the method of standard addition, a



                    linear curve  over the entire range of addition



                    is necessary  for the results to be considered



                    valid.




5.   Data Reporting




          Report all metal concentrations as follows:  Less than



     10 ug/1,  nearest yg; 10 yg/1 and above, two significant figures

-------
                                -  49  -
                            Cyanides

1.   Sample Preparation

          All samples are to be distilled prior to determination  for

     total cyanides.  The distillation procedure given on page 43

     of "Methods for Chemical Analysis of Water and Wastes,  (1974)"

     (Reference 7)  is to be followed.

2.   Sample Procedure

          The procedure for total cyanides as given on pages 43-48

     of "Methods for Chemical Analysis of Water and Wastes,  (1974)"

     (Reference 7)  is to be followed.

3.   Quality Assurance

          a)    Initially, demonstrate quantitative recovery with

               each distillation-digestion apparatus by comparing

               distilled standards to non-distilled standards.

               Each day, distill at least one standard to confirm

               distillation efficiency and purity of reagents.

          b)    At least 15% of the cyanide analysis will consist

               of duplicate and spiked samples.   Quality control

               limits are to be established and confirmed as described

               in Chapter 6 of the "Analytical Quality Control

               Handbook" (Reference 9) .

4.   Reporting of_ Data
                                      •
          Report cyanide concentrations  as follows:   Less than

     1.0  mg/1,  nearest 0.01 mg;  1.0 mg/1 and above,  two significant

     figures.

-------
                                -  50  -



                             Phenols


1.   Sample Preparation
                                      c

          Distill all samples prior to determination of phenols.


     Use the procedure in "Standard Methods for the Examination of

     Water and Wastewater," 14th edition, 1975, p. 576 (Reference 10),


2.   Procedure   :


          Use method 510 for phenols in Appendix Xf pages 577-580

     and 580-581.  Use method 510B for samples that contain less

     than 1 mg/1 of phenol.  Use method 510C for samples that contain

     more than 1 mg/1 of phenol.


3.   Quality Assurance


          Demonstrate quantitative recovery with each distillation

     apparatus by comparing distilled standards to non-distilled

     standards.  Each day distill, at least, one standard to con-

     firm the distillation efficiency and purity of reagents.


          Run duplicate and dosed sample analyses on at least 15%

     of the samples analyzed for phenol.  Establish and confirm

     quality control limits as described in Reference 9.


4.   Reporting of Data


          Report phenol concentrations as follows:

          Method 510B to the nearest yg/1.

          Method 510C - when less than 1.0 yg/1 to the nearest
                                      •
     0.01 mg; 1.0 mg/1 and above to two significent figures.

          Report all quality control data when reporting results

     of sample analysis.

-------
                                -  51  -
                           REFERENCES

1.   Determining Volatile Organics at Microgram-per-Liter  Levels
     by Gas Chromatography.  T. A. Bellar and J. J. Lichtenberg,
     Jour. AWWA, p. 739-744, Dec. 1974.
2.   Reference Compound to Calibrate Ion Abundance Measurements
     in Gas Chromatography—Mass Spectrometry Systems. J.  W.
     Eichelberger, L. E. Harris and W. L. Budde, Anal. Chem.  47,
     995-1000 (1975).
3.   ASTM Annual Standards - Water, part 31, Method D2908  "Standard
     Recommended Practice for Measuring Water by Aqueous-Injection
     Gas Chromatography."
4.   ASTM Annual Standards - Water, part 31, Method D3371  "Tentative
     Method of Test for Nitriles in Aqueous Solution by Gas Liquid
     Chromatograph."
5.   Harris, L. E.,   Budde, W. L.   and
     Eichelberger, J. W., Anal. Chem.,   46,  1912
     (1974).  "Direct Analysis of Water Samples for Organic Pollu-
     tants with Gas Chromatography-Mass Spectrometry-"
6.   Federal Register,  Volume 38, number 125, part II, Appendix II,
     p. 17319, Friday,  June 29, 1975, "Determination of Organo-
     chlorine Pesticides in Industrial Effluents,"
7.   Methods for Chemical Analysis of Water and Wastes (1974).
     U.S.  Environmental Protection Agency,  Technology Transfer.
8.   Determining Selenium in Water,  Wastewater,  Sediment and Sludge
     by Flameless Atomic Absorption Spectroscopy.  T.  D. Martin and
     J. F. Kopp, Atomic Absorption Newsletter 14,  109-116  (1975).

-------
                              .  - 52 -




 9.    Handbook for Analytical Quality Control in Water and Waste-



      water Laboratories (1972).   U.S.  Environmental Protection



      Agency,  Technology Transfer.



10.    "Standard Methods for the Examination of Water and Waste-



      water ,"  14th edition, 1975.

-------
                                - 53  -
                            APPENDIX I
                       General  Information

Emulsions
     Limited work with several  categories of industrial effluents
covered by this study  (tanneries, petroleum, soap and detergent,
steam electric, pesticide) show that emulsions of widely differing
frustration factors are often encountered in the extraction pro-
cedure.  Samples that emulsify  at basic pH usually also emulsify
at acid pH.  There are two equally acceptable alternatives avail-
able for the purposes of this protocol:  break the emulsion or start
over with fresh sample and use  a continuous extractor, to prevent
the formation of emulsions.
     By the 85% solvent recovery criteria, no way was found to break
the emulsion formed on extraction of  untreated tannery wastes.  A
soap and detergent sample was also very difficult.  The use of a
continuous heavier-than-water liquid  extractor allowed the extraction
to take place with no difficulties and very little labor.  However,
two days time is required.  Comparison of samples from four industries-
petroleum, tannery, pesticide, and soap and detergent—by both shake-
out and continuous extraction using wastes spiked with priority pollu-
tants indicate that the two techniques are comparable.  For some
individual cases one technique is better than the other but no clear
pattern emerges.  Therefore,  if desired, a continuous extraction
technique may be used in place of sepa-ratory funnel extraction for
all samples as well as those  for which it is absolutely necessary
because of intractable emulsions.

-------
                               -  54  -




                            APPENDIX I



                            (continued)




     There is a justifiable concern  that the extraction efficiency



for these compounds may differ widely depending on the nature of



the effluents.  This is true but no  better approach is apparent.



For example, recoveries of most of the base-neutrals were judged



to be about 75% from the tannery and petroleum samples but less



than 25% from soap and detergent.




Acid (Phenol) Analysis




     Although the 11 phenols of interest here do chromatograph on



the Tenax column cited, the chromatography is poor, particularly



for the nitrophenols.  Two other columns have shown good response



for the acid extractables.  SP2250 can be used for this purpose.



Phenol responses on SP2250 are shown in Table IV.  It should be



noted, however, that 4-nitrophenol,  2,4-dinitrophenol, 4,6-dinitro-



o-cresol, and pentachlorophenol failed to give MS response at the



100 ng level using this column.




     SP1000  (4% load) has also been  evaluated for use with the



acid fraction.  All but 2,4-dinitrophenol and 4,6-dinitrc-o-cresol



elute from this column.  Pentachlorophenol and 4-nitrophenol are



eluted from SP1000, but they produce broad peaks which are difficult



to quantify.

-------
                               - 55  -


                           Appendix II

      Possible Sources  for Some Priority  Pollutant  Standards
                                                      Source  of
Compound	Standard 2

acenaphthene                                          AN p.  118
acrolein                                              AL p.  18
acrylonitrile                                         AL p.  19
aldrin                                              HERL #30
dieldrin                                            HERL §2380
benzene                                                B p.  154
benzidine1                                          "  "  RFR
carbon tetrachloride  (tetrachloromethane)              B p.  88
chlordane  (technical mixture & metabolites)         HERL f!200

Chlorinated benzenes  (other than dichlorobensenes)

chlorobenzene                                         AL p.  165
1,2,4-trichlorobenzene                                AL p.  710
hexachlorobenzene                                     AL p.  416

Chlorinated ethanes  (including 1,2-
  dichloroethane, 1,1 /1-trichloroethane
  and hexachloroethane)

1,2-dichloroethane                       .             AL p. 261
1,1,1-trichloroethane                                  B p. 309
hexachloroethane                                      AL p. 416
1,1-dichloroethane                                '    PB p. 142
1,1,2-trichloroethane                                 PB p. 388
1,1,2,2-tetrachloroethane                             'PB p. 372
chloroethane                                          EA p. 53

Chloroalkyl ethers (chloromethyl, chloroethyl and
  mixed ethers)

bis(chloromethyl)  ether1                                EFRi
bis(2-chloroethyl)  ether                              AL p. 173
2-chloroethyl vinyl ether                             AL p. 174

Chlorinated naphthalene
                                           •
2-chloronaphthalene                                  ICN p. 50

-------
                             - 56 -

                           Appendix II

      Possible Sources for Some  Priority Pollutant Standards
                            (Continued)
Compound
 Source of
 Standard2
Chlorinated phenols  (other than those listed
  elsewhere;includes trichlorophenols and
  chlorinated cresols)
2,4,6-trichlorophenol
p-chloro-m-cresol
chloroform  (trichloromethane)
2-chlorophenol

DDT and, metabolites

4,4'-DDT
4,4'-DDE
4,4f-DDD  (p,p'-TDE)
                                    V
Dichlorpbenzenes  (1,2-;1,3-; and 1,4-
  dichlorobenzenes)

1,2-dichlorobenzene
1,3-dichlorobenzene
1,4-dichlorobenzene

Dichlorobenzidine

3,3*-dichlorobenzidine*

Dichloroethylenes  (1,1-dichloroethylene and
  1,2-dichloroethylene)

1,1-dichloroethylene
1,2-trans-dichloroethylene
2,4-dichlorophenol

Dichloropropane and dichloropropene

1,2-dichloropropane
1,3-dichloropropylene  (1,3-dichloropropene)
2,4-dimethylphenol

Dinitrotoluene

2 ,4-dinitrotoluene
2,6-dinitrotoluene
1,2-diphenylhydrazine
  AL p.  712
 TCI p.  102
   B p.  92
  AL p.  187
HERL 51920
HERL *1860
HERL 11780
  AL p.  258
  AL p.  258
  AL p.  258
 CPL p.  81
  AL p.  746
  AL p.  262
  AL p.  265
  AL p.  267
  AL p.  267
  AL p.  323
  PB p.  180
  PB p.  180
  AL p.  338-

-------
                             - 57 -

                            Appendix II

      Possible  Sources  for Some  Priority Pollutant Standards
                            (Continued)
Compound
 Source of
 Standard 2
Endosulfan and metabolites

ct-endosulfan
g-endosulfan
endosulfan sulfate

Endrin and metabolites

endrin
endrin aldehyde
 HERL .£3220
 HERL §3200
   NI p.  45
 HERL §3260
  NI p.  147
ethylbenzene
fluoranthene

Haloethers  (other than those  listed elsewhere)

4-chlorophenyl phenyl ether  (p-chloro-
  diphenyl ether)
4-bromophenyl phenyl ether
bis(2-chloroisopropyl) ether
bis(2-chloroethoxy) methane

Halomethanes  (other than those listed elsewhere)

methylene chloride  (dichloromethane)
methyl chloride  (chloromethane)
methyl bormide (bromomethane)
bromoform (tribromomethane)
dichlorobromomethane
trichlorofluoromethane
dichlorodifluoromethane
chlorodibromomethane

Heptachlor and metabolites

heptachlor
heptachlor epoxide
hexachlorobutadiene

Hexachlorobyclohexane (all isomers)

cx-BHC
6-BHC
y-BHC (lindane)
6-BHC
   B  p.
  AN  p.
161
118
 RFR p.  6*
 ICN p.  37
  PB
  PB p.  62
PB
PB
PB
PB
CO
PB
PB
CO
P-
P-
P.
P.
P.
P.
P.
P.
276
277
276
73
16
358
142
27
HERL #3860
HERL #3880
  AL p. 416
HERL #620
HERL #640
HERL #680
HERL #660

-------
                               - 58 -

                           Appendix  n

      Possible Sources for Some Priority Pollutant Standards
                            (Continued)
Compound
 Source of
 Standard 2
hexachlorocyclopentadiene
isophororie
naphthalene
nitrobenzene

Nitrophenols  (including 2,4-dinitrophenol and
  dinitrocresol)

2-nitrophenol
4-nitrophenol
2,4-dinitrophenol
4,6-dinitro-o-cresol

Nitrosamines

N-nitrosodimethylamine*
N-nitrosodi-n-propylamine
N-nitrosodiphenylamine
pentachlorophenol
phenol

Phthalate esters                            j

bis (2-ethylhexyl) phthalate
butyl benzyl  phthalate
di-n-butyl phthalate
diethyl phthalate
dimethyl phthalate

Polychlorinated biphenyls  (PCB's)

PCB-1242  (Arochlor 1242)
PCB-1254  (Arochlor 1254)

Pqlynuclear aromatic hydrocarbons  (including
  benzanthracenes, benzopyrenes, benzo-
  fluoranthene, chrysenes, dibenzanthracenes,
  and indenopyrenes)

1,2-benzanthracene
benzo [a]pyrene  (3,4-benzopyrene)
3 r4-benzofluoranthene
11,12-benzofluoranthene
chrysene
  AL p. 416
  AL p. 464
  AN p. 118
  AL p. 557
  AL p. 564
  AL p. 564
  AL p. 332
 TCI p. 188
  NI p. 173
  PB p. 310
  EA p. 159
  AL p. 587
  AL p. 595
  CS p,
  CS p.
  CS p.
  CS p,
  CS p.
8
8
8
8
8
HEKL #5703
HERL 15705
  AN p.
  AN p.
  NI
  NI
  AN p.
118
118
118

-------
                              - 59 -

                           Appendix n

      Possible Sources for Some Priority Pollutant Standards
                           (Continued)
Compound
2,3,7,8-tetrachlorodibenzo-p-dioxin (TCDD)
tetrachloroethylene
toluene
toxaphene
trichloroethylene
vinyl chloride (chloroethylene)
1-bromodecane (possible internal standard)
1-bromododecane (possible internal standard)
Source of
Standard 2
acenaphthylene
anthracene
1 , 12-benzoperylene
fluorene
phenanthrene
1,2:5 , 6-dibenzanthracene
indeno (1,2 , 3-C,D) pyrene
pyrene
AN p.
AN p.
AN p.
AN p.
AN p.
AN p.
AN p.
AN p.
1
118
118
118
118
118
118
118
  NI p. 174
  AL p. 680
  AL p. 701
HERL #6740
  AL p. 711
  PB p. 406
Footnotes :
    These compounds or any mixture containing 1% or more by weight
    of these compounds are defined as carcinogens in the Federal
    Register, Vol.  38, No. 144,  pp.  20074-20076, 27 July 1973.
    Prescribed safety regulations for handling are in the Federal
    Register, Vol.  39, No. 20,  pp. 3756-3797, 29 January 1-974.
                                                     i
    Only one source is listed even though several may be available.
    These sources are not to be  interpreted as being endorsed by
    the EPA; they serve to show at least one ven--'.or where each
    standard can be obtained.  When  several sources were available
    and compound purity was listed,  the source having the highest
    purity material was selected.

    These compounds have been ordered but have not been received
    at Athens ERL as yet.

-------
                                - 60  -


                Sources  of Standards and Abbreviations



  AL      Aldrich  Chemical Co., Milwaukee, Wise.;  Catalog 1977-1978.

  AN      Analabs,  Inc., North Haven, Conn.; Catalog 18  (June 1976).

   B      J. T. Baker Chemical Co., Phillipsburgh, N.J.;
          Catalog  750 (July 1975).

  CS      Chem-Service,  West Chester, Pa.; Bulletin  CS-100-8 (1975).

 CPL      Chemical Procurement Laboratories, College Point,  N.Y.;
          1975 catalog.

  EA      Eastman  Kodak  Co., Rochester, N.Y.; Catalog 48  (1976).

 ICN      K&K Rare &  Fine Chemicals, Plainview, N.Y.; Catalog No. 10
           (1975).

  NI      Nanogens International, P.O. Box 487, Freedom, CA   95019
          "Nanogen Index" (1975).

  PB      Pfaltz & Bauer Chemical Co., STamford, Conn.; Catalog
          1976.

 RFR      RFR Corp.,  Hope, R.I.;  "Chemical Standards for  Air-Watsr-
          Industry-Foods" (1975).

HERL      "Analytical Reference Standards and Supplemental Data for
          Pesticides  and Other Selected Organic Compounds",  EPA-
          660/9-76-012 (May 1976),  Health Effects  Research Laboratory,
          Environmental  Toxicology Division, Research Triangle Park,
          NC.  A sample  order blank for standards  and the above  •
          publication are attached.
                                                        t
  CO      Columbia Organics Catalog A-7, Columbia, S.C.   (1975).

 TCI      Tridom Chemical Inc., Haut-tauge, N.Y. , Catalog  No. 1
           (1976).
               *Those labs which are under contract to perform GC-MS analysis
                for EPA, may obtain a set of standards from *fr. William A.
                Telliard, Chief, Energy and Mining Branch, Effluent
                Guidelines Division, (WH 552) 401 M Street, S.W.
                Washington, D.C.  20460  (202) 426-2720.

-------
                                      -  61 -
                         ENVIRONMENTAL TOXICOLOGY DIVISION
                         HEALTH EFFECTS RESEARCH LABORATORY
                  UNITED STATES ENVIRONMENTAL PROTECTION AGENCY
                   Research Triangle Park,  North Carolina  '27711
SUBJECT-  Index of Pesticides Analytical Reference
         Standards - Update of Hailing List
                                                         DATE:  June,  1976
FROM:
TO:
   Health Effects Research Laboratory, ETD,  ACB,
   Research Triangle Park, NC, U.S.A. 27711  (MD-69)
       l"
   All Laboratory Facilities on our Mailing  List

This copy of the 1976 revision of our pesticides  reference standards
index was mailed to the address appearing on our  mailing list.  As this
list is several years old, we are sure that  a number of addresses have
changed and that some are probably no longer existent.

If you wish to remain on our mailing list to receive future updates
of this publication, would you be good enough to'  complete the mail-back
below, snip it off,  and return it to us.   Do_ not  tear off the back cover
to return to the address shown.  If you have no use for this publication
but know of some other individual within  your organization who is con-
cerned with pesticides analysis,  would you convey this index, along with
the mailback,  to that person.	
 To:  U. S. Environmental Protection Agency
     Health Effects Research Laboratory
     Environmental Toxicology Division
     Research Triangle Park, NC, U.S.A. 27711
                                                             pate
                                          >-69)
    D
    D
    D
   We  wish  to be retained on your mailing list to receive future  updates
   of  the Pesticides Standards Index.  The address shown on the envelope
   is  entirely correct and requires no changes.
  We have no interest in future updates of this publication.
  cancel us from your mailing list.
Please
  We wish to be retained on your mailing list, but  the address shown
  on the envelope should be changed to read (print  or type)

-------
                                       -  62 -
                   REQUEST FOR ANALYTICAL REFERENCE STANDARDS
TO:  Quality Assurance Section
     Environmental Toxicology Division,
     EPA, HERL, Research Triangle Park, NC 27711
     MD-69
                                                                    Date
Date Request Reed
Shipment Date
Lab Code No.
Order Filled by
DO NOT WRITE IN THIS SPACE
The following reference standards are required for our program:
Catalog Compound H Catalog Compound
Code (Catalog Name) n Code (Catalog Name)
No. | No.

1
h
i
1







1





1



 If necessary,  use  back of  sheet  to  complete  list.  Covering letter unnecessary
 if this form is  completed  in full.
            Name and  address  of laboratory
                                       Requestor's Name  (Print or type)
 IMPORTANT:
   1.   The amount of each standard is  restricted to 100 mg because of the scarcity
       and expense of refining analytical  grade materials.

   2.   Please return at once, the  acknowledgement card enclosed with each shipment.
       This provides the sole evidence of  delivery of the shipment.

   3.   Do not request compounds not listed in  the catalog.  No others are stocked.

   4.   If a bottle appears to be  empty,  remove cap and examine interior of bottle
       and cap.   Certain highly viscous  materials tend to collect in cap.

-------
        .  ,             APPENDIX III


    ..."1"~ COLLECTION OF SAMPLES FOR SCREENING ANALYSES
The Initial  characterization  (screening)  of the  varied  industrial
discharges covered  by  this  program vn"!l be  made  on  an analysis' of
a composite  effluent sample.  Any scheme  for collecting a
composite sample  is, in  effect  a method for mechanically
Integrating  to  obtain  average characteristics of a  discharge.
During the screening phase  the  sample composite  can be  used to
determine the average  characteristics which would be
representative  of that discharge.  Simple composite samples are
those that are  made up of a series of aliquots of constant volume
collected at regular time intervals in a  single  container.  Some
situations may  require flew or  time proportional sampling, this
determination will be  made by the individual project officer
after considering his  specific  industrial category.

The determination of compositing period 24,  48 or 72 hours will
be made on a case by case basis.  The duration of compositing
will depend  on  the type  of sample being collected,  the type of
facility being  sampled and the  time varying  characteristics of
the discharge.  The rate of change of flow  and other
characteristics of the discharge and the  accuracy required will
also influence  tha determination of the compositing period. For
example longer  compositing periods would  be warranted when less
•stable unit  process    operations are being sampled.

Collection of Samples

1.  Collection  of Composite Samples for Liquid-Liquid Extraction

    Collect  a representative composite sample.  The maximum time
    interval between aliquot samples shall be no longer than 30
    minutes.  The minimum aliquot size shall be ICO ml. •/ The
    sample must be collected with an automatic sampler using the
    equipment and methods outlined below.   Minimum composite
    volume must be 2 1/2 gallons. -^'

    Automatic Sample Collection

         Sampler - A peristaltic pump automatic sampler with  ,
    timer and a single glass compositing  jug is  required.   The'2
    1/2 - 3 gallon compositing bottle must be glass  and  cleaned

-------
                          -  64 -
  as  outlined below.   New unused tubing  must be  used  for the
  sampling line and for the pump for each  individual  outfall  or
  sample location.   Vacuum type automatic  samplers  may  be used
  provided that the sample chambers  are  glass and  that  they are
/ cleaned after every use as outlined for  glass  composite
  containers.  Place  the sampler or  composite container in an
  Insulated chest and ice.  Maintain the sample  at  4°C  during
  the compositing procedure.  At the completion  of  the
  compositing period  seal the container  with a teflon lined
  cap.   Place the container in an. insulated  shipping  container,
  1ce,  and seal, then ship to the analytical  laboratory.
t Maintain at 4"C during transport and storage prior  to
'' analysis.

       When sampling  raw untreated industrial  discharges  which
  are generally high  in suspended solids it  is imperative that
  adequate sample flow rate be maintained  throughout  the  sample
  train in order to effectively transport  the solids.   In
  horizontal  runs,  the velocity must exceed  the  scour velocity,
  while in vertical runs the settling or the fall velocity must
  be  exceeded several times to assure adequate transport  of
  solids in the flow.  The equipment used  in sampling raw
  discharges  than must have a minimum intake velocity of  2 feet
  per second.  In the sampling of treated  effluents just  about
  any commerically available automatic liquid sampler could be
  used.

       When more than one laboratory is  involved in the
  analysis of the various parameters, the  sample should if at
  all possible not be divided in the field but rather at  the
  contractors' laboratory.  For purpose  of this  program the
  composite will be divided into four parts,  one part for
  metals analysis,  one for pesticides and  PCB's, one  for  GC/MS
  compounds and one for the classic  parameters.

       Blend the composite sample to provide  a homogeneous
  mixture including a representative suspension  of any  solids
  in  the container.  No specific method  is required,  hand
  stirring with clean glass or teflon rods,  mechanical  paddles
  or  magnetic mixing  with teflon coated  stirring bars may be
  used.  Metal mixing devices may not be used.

       Metals - Withdraw a well  blended  aliquot  of the
  composite sample.  Using a glass funnel, rinse the sample
  container with a  small  portion of  the  sample,  then  transfer

-------
                         -  65 -
250 - 500 mcf/1 of sample to  the bottle.   Do  not  add  any
preservative to the sample just seal  and  prepare for
Shipment.  AlT~samples must  be carefully  identified  using
labeles supplied by EGO.  Indicate on the  label  whether the
sample is a raw discharge or treated effluent as  shown.  If
sample is to be run on the plasma unit only  indicate so  at
base of tag.  Ship samples to the Chicago  Regional Laboratory
at the addressed shown.
            U.S. Environmental Protection Agency
            Region Y, Central Regional Laboratory
            1819 W. Pershing Road
            Chicago, Illinois   60609
              Raw discharge or treated effluent


                            N2 V002200

                 Location	

                 Sampler	
                Sample Point.
                Timm
                         rf^/

                Plasma only

-------
                            - 66  -
Field Blank Procedure for Automatic Samplers

    Blank Water - Blank v/ater must be as free from organic
Interferences as possible.  The analytical laboratory should
supply this water in bulk glass containers (m'nimum of five
liters) for field use.  The supplying laboratory shall analyze
the blank water to determine the organic background that may be
present.

    Procedure - All parts of the sampling system must be scrubbed
with hot detergent water and thoroughly rinsed with tap v/ater and
blank v/ater prior to use.  Further rinsing with methylene
chloride is required when parts permit, i.e., are"not susceptible
to dissolution by the solvent.  (Note: Tygon plastic tubing is a
source of ph thai ate ester contanii nation.  Where its use is
required, i.e., in the peristaltic pump, the length must be kept
as short as possible.  Teflon is acceptable and may be used in
other parts of the sampling system as in intake lines.  In the
field, pump two liters of blank water through the sampling line
and pump tubing and discard.  Then pump three liters of blank
water through the system and collect as a blank in a 1-gallon
sample bottle that has been prepared as described below.  Seal
the bottle with a Teflon lined cap.  Immediately ice the blank (4°
C) and maintain at (4°C) during the transport and storage prior
to analysis.

    Composite Container - Prepare narrow-mouth glass sample
bottles for use by washing with hot detergent water and
thoroughly rinsing with tap v/ater and blank water.   Heat the
bottles at 4CO°C in a muffle-furnace or dry heat sterilizer for
30 minutes or alternatively, rinse with methylene chloride and
air dry .at_rqonf>empefature protected from atcrnspheric or other
sources of contamination.  Caps for the bottles must be lined
with Teflon which has been solvent rinsed as above.

2.  Collection of Grab Samples

         Collect grab samples ( minimum of one per day) for the
    analysis" of phenol, cyanide, and volatile organics
    (purgable).  Collect samples from the raw process discharge,
    the treated effluent, and the treated effluent after
    chlorination, when chlorination is practiced.   It is
    recommended  that the samples be collected from miji'^chajinel
    at nrfd-dep_th.  Samples should  be  collected  at a turbulent,
    well mixed section of the channel.

-------
                             - 67  -
    Cyanide (Total)

         Container - Use nev/ one-liter plastic bottles that will
    not contaminate the sample.  Wash the bottles and caps with
    hot water and thoroughly rinse with tap water and blank
    water.

         Collect a 1-Hter sample.

         Pretreatment and Preservation - Oxidizing agents
    such as chlorine decompose  many  cyanides.  Therefore, at
    time of collection, samples must be treated to eliminate
    such agents.  Test a drop of the sample at the time of
    collection with potassium iodide-starch test paper
    (Kl-starch paper);  a blue color  indicates the need for
    treatment.  Add ascrobic acid, a few crystals at a time,
    until a drop of the sample  produces no color on the indicator
    paper.  Then add an additional 0.6 g of ascorbic acid for
    each liter of sample volume.  Then add 2 ml of 10 N sodium
    hydroxide per liter of sample  (pH >_ 12).

         Seal the sample bottle and place in an insulated chest
    and tee (4*C).  Seal the chest and ship to the analytical
    laboratory.  Maintain at 4'C during transport and'storage
    keep out of direct light prior to analysis.

Phenols

    Container - Use new one-liter glass bottles.  Wash the bottle
and Teflon cap liner with hot water and thoroughly rinse with tap
water and blank water.                                         1

    Collect a 1-liter sample.

    Preservation - At the time of collection, acidify the sample
by addition of phospheric acid or sulfuric to pH 4.  Note volume
of acid added on sample tag.   Seal bottle, place in insulated  -
chest and ice (4"C).  Seal chest and'ship to analytical
laboratory.  Maintain at 4°C during transport and storage.  Keep
out of direct light prior to analysis.

-------
                            -  68 -
         Organlcs (Purge and. Trap Method)

    Containers - Use 45 to 125  ml   screw cap glass vials with
Teflon faced si 1 cone septa:     '

         Vialsfa)- Pierce #13074 or equivalent

         Septafa)- Pierce #12722 or equivalent

    Wash the bottles, septa, and caps with hot water and
thoroughly rinse with tap water and blank water.  Keat the
bottles and septa at 105°C for one hour, cool to room temperature
1n an enclosed contaminant free area.  When cool, seal bottles
with septa (Teflon side down) and screw cap.  Maintain the
bottles in this condition until just prior to fillinc with blank
water or sample.

         Available from Pierce, Inc. Box 117, Rockford, IL 61105.
    Collect duplicates 45-125 ml samples each time samples are
collected.  Two blank water samples, sealed in 45 ml vials, are
to accompany the sample bottles curing shipment to and from the
sampling site.  If preservation for residual chlorine is to be
used, collect four samples during each sampling period.  Two
should be preserved and two not preserved.  Two preserved and two
non-preserved blanks are to be provided.

    Filling and Sealing Bottles - Slowly fill each container to
overflowing.  Carefully set the container on a level surface.
Place the septum (Teflon side down) on the convex sample
meniscus.  Seal the sample with the screw cap.  To insure that
the sample has been properly sealed, invert the sample and
lightly tap the lid on a solid surface.  The absence of entrapped
air bubbles indicates a proper seal.  If air bubbles are present,
open the bottle, add additional sample, and reseal.   The sample
must remain hermetically sealed until it is analyzed.

    Preservation - Preservative (sodium thiosulfate  or sodium
bisulfite) is used to stabilize samples containing residual
chlorine.  The production of chloroform and other haloforms
continues in such samples if they are not stabilized.  Waste
streams that have been treated with chlorine should  be tested on

-------
                             - 69  -
site to determine whether or not preservative is  needed.   If
preservetation is required,  collect both  preserved  and  non-
preserved samples.  Wrap the samples with water proof packing
material, place in an insulated chest and ice at  4°C.   Maintain
at 4*C during transport and  storage prior to  analysis.

3.  Identification of Samples

         All samples and blanks must be carefully identified
    using water proof labels and water proof  ink.   Include the
    following information on the label: sample  number,  date  and
    hour of sampling, complete  information as  to  source and
    sampling point, preservative added, if any, and name of
    person collecting the sample (include address and/or phone
    number).

-------
APPENDIX IV

-------
                  REFERENCE NO. 1
DETERMINING VOLATILE ORGANICS AT MICROGRAM-PER LITER LEVELS

-------
 ;^V;-^;:'?^ --:^".:^'i
DECEMBER 1974
                                           Determining
                                           Volatile Organics
                                           atMicrogram-
                                           per-Litre Levels
                                           by Gas
                                           Chromatography

                                           T. A. Be/far and J. J. Lichtenberg
                                                                A Metricized Article
                                           A contribution submitted to the JOURNAL on Nov. 7, 1974, by
                                           T. A. Bellar and J. J. Lichtenberg, res. scientists. Methods Oev.
                                           and Qual. Assurance Res. Lab., EPA, Natl. Envir. Res. Ctr, Cin-
                                           cinnati, Ohio.
Presented here is a method for quantitative recovery
of volatile organic compounds followed by a
description of apparatus and procedures employed to
detect 0.5 /xg/l of the substances.


  Recent legislation1-2 requires analytical methods for the
determination of hydrocarbon and chlorinated organic sol-
vents in wastewater. In some cases a minimum detectable
limit of 1 /ig/I (10~3ppm) is required for specific compounds.
It is the responsibility of the EPA's Methods Dev. and Qual.
Assurance Res. Lab. to evaluate existing methods, and when
necessary, to develop new methods to meet such needs.
  Determination of these substances at the 1-p.g/l (lO'-'ppm)
level has been difficult. Commonly used techniques such as
direct aqueous-injection gas chromatography,  liquid-liquid
extraction, and head-gas analysis have proved inadequate.
Direct aqueous-injection gas chromatography3- * although
generally useful for analysis of industrial effluents, provides
an approximate limit of detection of only 1 000 ^ig/l (1 ppm).
  Liquid-liquid extraction methods using low5 or high6 boiling
organic solvents followed by gas-chromatographic analyses
have been  investigated. These methods have provided erratic
or low extraction efficiencies for volatile compounds. In addi-
tion large solvent responses and solvent impurities can cause
serious  chromatographic interferences. Distillation  tech-
niques7 have been employed in which a small quantity of sam
pie distillate is collected and analyzed by direct aqueous-injec-
tion gas chromatography. Detection limits of approximately I
/ig/1 (10~J ppm)  have been reported for water-soluble
volatiles using this method. Poor recoveries render the meth-

           T. A. BELLAR AND J. J. LICHTENBERG  739

-------
                         TABLE i
                    Trap-Saturation Volumes
Compound
Met hint
Ethane
Propane
n-Buune
a-Pentane
a-Huine
D -Aitanes
CT-CIS
Benzene
Touione
Methylene
chloride
Chloroform
Aldehydes
C2 and above
Phenols
Naphthylene
Chloro benzene
o-Dichlorooenzene
1.2.4 Tncbloro-
benzene
Silica
Gel
Layer
ml
<5'
<25'
>50'
>500'
>500-
>5W

t
>5oo-
>500*

«
<
Nonquanu-
tau e






Pore Pak
Q
ml
<5"
<5'
<50'
<100'
<250t
>500T

>500t
>500*
>500t

>55W

>500-
>500-
>500'
t
t
t

Chromo-
sorb
103
ml
<5'
500t

>500T
>500t
>soot

>50W
>SW

>500*
>soo-
>550W
>50W
>500«

Tenax
GC
ml
' T
t
<
t
t
>soot

>500T
>500T
>soot

>5005
>55W
>500*
>50M

Retenuon
Index
100
200
300
400
500
600

700-1 500
_
_

_
_

_
_
_
_
_
_

•Values reported by Bellas and Sigsby '
rvalues determined using water-saturated nitrogen as purge gas are same as those reported
under dry conditions.
• Not determined
{Determined Tor this study


                          TABLE 2
            Purging of Selected Compounds From Water
    Nitrogen Purge Gu
                          Percentage Remaining in Aqueous Phase
Bow Rate
mlAntit
20
20
20
20
13
13
13
13




Volume
ml
0
20
100
300
0
6J
85
143




Methylene
Chloride
100
60
0
0
100
67
30
6

2

•J0.1
Chloroform
100
55
0
0
100
94
29
0
Solubility in w
1
Boiling p
61.3
Benzene
100
46
3
0
100
71
6
0
net— per ant
0.08
»ni— C
80.1
2-Buunone
100
95
%
80
100
100
36
74

35

79.6
                          TABLE 3
              System Response to Methylene Chloride
Concentrauon
u.X/1
5.2
104
20.8
52.0
104.0
260.0
5204)


Slope
0.0 to Data Point
78.5
76.9
78.5
77j
76J
77.9
85.7
78.0 mean
3.88 sid. dev.
Diluuon
1/100
2/100
5/100
10/100
20/100
50/100
Stock solution


od useless for water-insoluble components.
  The head-gas technique* in which  a sample is sealed in a
partially filled container, has been employed for many years.
Each volatile organic compound establishes  an  equilibrium
between the gaseous and  aqueous phase. At  low concentra-
tions the ratio of the concentration in the gaseous phase to the
concentration in the aqueous phase  is  a  constant  (partition
coefficient) and is unique for each  organic  compound. By
analyzing the gaseous phase and applying the appropriate par-
tition coefficient, one can calculate the concentration of each
organic initially present in the aqueous  phase.
  Of the  techniques previously  mentioned, the  head-gas
method has the greatest potential for meeting the  needs set
                                                                                          TABLE 4
                                                                                 System Response to Chloroform
Concentration
M//
6.2
12.4
24.8
62.0
124.0
310
620


Slope
0. 0 to Data Point
32J
29.7
28.1
26.8
26.8
25.4
29.9
28.4 mean
2.35 »d. dev.
Dilution
1/100
2/100
5/100
10/100
20/100
50/100
Stock solution


                                                                                         TABLE 5
                                                                                  System Response to Benzene
Concentration
Mf//
3.5
7.0
14.0
35.0
70.0
175
350


Slope
0, 0 to Data Point
220.6
219.4
214.9
215.8
207.5
196.0
232.0
215J mean
11.4 suLdev.
Dilution
1/100
27100
5/100
to/ioo
20/100
50/100
Slock solution


                                                                                         TABLE 6
                                                                                  System Response to Toluene
Concentration
u-gll
3J
70
14.0
35.0
70.0
175.0
350.0


Slope
0. 0 to Data Point
120
120
116
115
111
105
124
116 mean
6 .27 sid. dev.
Diluuon
1/100
2/100
5/100
10/100
20/100
50/100
Stock solution


                                                                                         TABLE 7
                                                                          Purging Efficiency at 19.5 C, Percentage Recovery

Purge
Volume
mINj
0-60
60-120
120-240
240-360
360-180
480-600
600-770
770-440
840-960
960-1080
1 080-1 200
Compound and Boiling Point

n-C5
36C
100











"-C6
69C
100











n-C7
we
98
2










n-Cg
126C
90
6
3
I








n-G,
HOC
76
12
8
3
I







n-C,,
I96C
60
15
9
4
2
2
2
I
1
1
2

n-C,3-
.'.WC
44
17
13
6
3
5
3
2
2
2
2

n-C15'
270C
2
13
27
14
3
7
5
5
4
4
7
                                                                •Not 100 per cent purged using 1 560 ml N.
                                                                                         TABLE 8
                                                                          Purging Efficiency at 65C, Percentage Recovery
Purge
Volume
ml Nj
0-60
60-120
120-240
240-360
360-480
480-600
600-::o
770-840
Compound and Boiling Point
n-Cj
J6C
100







n-C6
69C
100







n-C7
100







s&
100







n-Cj
HOC
97
3






n-Cn
/96C
76
10
6
4
2
1
1

n-Cu-
.'J-JC
66
12
6
6
4
3
2 ^
1
•Sf
27
24
• 15
11
7
6
r 6
4
                                                                'Not 100 per cent purged
740   RESEARCH
                                                                                                         JOURNAL AWWA

-------
              !)•»- l/4-.n.-OD Inltt
 1/4-m.-00 Exit
      10-mm Gin
      Mtdium Poresitv
                       Simon Inltt
                       S-mm-OO Rubber Swtum
                         10-mm 00
                           - l/4-in.-OO Inltt


3odv t
Aavnbly —^f~
Conntctinq - m | f
Nua X^S
Stem £
Awmbly — »W















1
^ Tno V«nt
I
5
3
ii
^ cm

• Glw-Wool Plug ~Z 16
	 15
^ ^
— 13
— 12
50-«OMwn 	 11
Porous Polvmtr Sttds 	 \Q
-^ 9
— 3
	 7
5 Ptf Cjnt Hiqn —
T-motraturB ciastomtr- ~— 4
200 on SO-80 Mtin _Z 3
Chromo*oro-G ~ -
— 2
3l«»-
-------
maximum bubble contact time and efficient mixing.
  Gaseous volumes above  the sample reservoir are kept to a
minimum to provide efficient transfer characteristics and
allow sufficient space in which most foams can disperse. Inlet
and  exit ports are constructed from  0.06-mm (l/4-in.)-OD
medium- or heavy-wall  tubing so that leak-free  removable
connections can be made  using "finger-tight" compression
fittings containing plastic ferrules. The optional foam  trap is
used to control occasional samples that foam excessively.
   Trap. The trap (Fig. 2) is a short section of stainless-steel
tubing packed with an adsorptive material such as gas-chro-
matographic grade porous  polymers, silica gel, or molecular
sieve.  Volatile materials are transported directly from the
purging device into the trap by the purge gas. The adsorbent
retards the flow of the purged compounds while the purge gas
is  vented. The properties of the adsorptive material are cho-
sen to meet the needs of the particular analysis. The following
criteria must be met.
  The volume of the purge gas passing through the adsorbent
packed in the trap can approach but not reach the retention
volume of the compound to be trapped (See Table I).
  The retained  compounds must not be irreversibly sorbed
by the  trap. (Silica gel irreversibly adsorbs some aromatics
above C?).9
   No chemical  reactions or rearrangements may take place
when the sample is being concentrated, stored, or desorbed.
(Silica gel causes externally bonded olefins to rearrange to the
cis- and trans-2 olefins).^
   The  adsorptive material must be  thermally stable.
Chromosorb 103 and Tenax GC have been found to perform
satisfactorily. (Divinyl benzene crosslinked porous polymers
out-gas extraneous compounds causing serious interferences
during mostgas-chromatographic analyses.)10
   The trap is assembled and packed with the appropriate ad-
sorptive material according to Fig. 2. The body assembly acts
as a seal for the exit end of the trap. The modified stem assem-
bly is used to attach the trap to the desorption device. The cap
is used to seal the inlet end of the trap when it is not  in use.
   Desorbers. The desorbers (Fig. 3, 4)  are used to transfer the
contents of the  trap  to  the gas chromatograph for analysis.
This is done with the use of an auxiliary carrier flow-control
system which back/lushes the  trap at elevated temperatures
directly onto the gas-chromatographic column. Desorber 1 is
used exclusively with one type  of gas chromatograph, but
desorber 2  can  be used as a universal desorber for  all gas
chromatographs with a septum-type liquid-inlet system.
   Desorber 1 (Fig. 3) is attached directly onto the gas-chro-
matograph liquid-inlet system after removing the septum nut,
the septum, and the internal injector parts. The modified body
assembly is screwed onto the inlet system using the TFE
gasket as a seal. A plug is attached to one of the stem assem-
blies. The assembled parts, simply called "the plug," are used
to seal the desorber whenever the trap is removed to maintain
the  flow of carrier gas through the gas-chromatographic col-
umn. The flow controller, TFE tubing, and stem assembly are
used to provide the trap-backflush flow. This entire assembly
is also used to provide gas flow to operate the purging device.
   Desorber 2 (Fig. 4) may be attached to any gas chroma-
tograph by piercing :he  GC liquid-inlet septum with the  nee-
dle. The desorber is assembled according to Fig. 4 with inter-
nal volumes and dead-volume areas held to a minimum. The
heat source is concentrated near the base of the desorber so
 that the internal seals of the body assembly do not become
damaged by heat. Tine use of a detachable needle assembly
from a microsyringe makes it easy to replace plugged or dulled
needles. The flow controller, TFE tubing, and stem assembly
are used  to provide  the trap-backflush flow.  This entire
assembly is also used to provide gas flow to operate the purg-
ing device.
    A. gas chromatograph was equipped with  dual-flame
ionization detectors and a microculometric  detector (halide
mode).
  Column 1 consisted of dual, stainless-steel, 180-cm (6-ft)
long  x 2.67-mm (0.105-in) ID  columns, packed with
Chromosorb-101 (60/80 mesh). The carrier gas was nitrogen
at 50 ml/min (O.cu ft/hr). The oven temperature was isother-
mal 190C (310F) or programmed from 120C to 225C (247F to
437F) at 10C(50F)/min.
  Column 2 consisted of dual, stainless-steel, 91-cm (3 ft) x
1.65-mm-  (0.065-in.)- ID columns packed with 4  per cent
SE-30 on Chromosorb-P (NAW) (60/80 mesh). The carrier
gas was nitrogen at 50 ml/min (0.1  cu ft/hr). The oven tem-
perature was programmed from 60C to 230C (140F  to 446F)
at 10C(50F)/min.
  The GC-MS system consisted of a  gas chromatograph'
with a mass spectrometer! controlled by a data-acquisition
system.* The column was glass, 240-cm (8-ft) long  x 2-mm
(0.078-in.)  ID  and packed  with  Chromosorb-101 (50/60
mesh). The carrier gas was helium  at 30 ml/min (0.06 cu ft/
hr). The initial oven temperature was 125C (257F) for 3 min
and then programmed at 4C  (39F)/min to 220C  (428F).

Reagents
  Organic-free  water was prepared  by passing distilled water
through a water-treatment system.§
  Standard stock solutions were prepared by injecting 1-5 p.\
61.02  x lO^cu. in. of the compound to be determined into a
1-1  (61-cu in.)  volumetric flask partially filled with  organic-
free water. The mixture  was then diluted to volume with
organic-free water to give concentrations between  1 and 7
mg/1 (1 and 7 ppm). Dilutions were made from the stock solu-
tion by pipetting a known quantity of stock solution into a
partially filled volumetric flask and diluting to volume with
organic-free water.  [For  low-level  work  (1-10  /ig/0
(10~3-10'2ppm) a 1:10 dilution of the stock solution  was pre-
pared, and secondary dilutions were prepared.]
Procedure
  Purging and trapping. With nitrogen gas flowing through the
purging device (Fig. 1)  at 20  ml/min (0.04 cu ft/hr), the trap
inlet (Fig. 2) was attached (finger-tight) directly to the purg-
ing device exit using a compression  fitting. The trap vent was
inserted into the exit end of the trap. Five millilitres of sample
were  injected  into the purging device and purged for the
specified time (11 min). The trap was then removed  from the
purging device, and the vent plug was removed and  replaced
with a cap to seal the trap inlet.
  Trap conditioning. Newly packed traps were conditioned at
approximately 200C (392F) with a nitrogen flow of 20, ml/min
(0.04  cu ft/hr)  for 16-24 hr  with one of the desorbers and
vented to the room. Prior to daily use, traps were placed into
the desorber and conditioned at 130C  (266F)  for  approxi-
mately 10 min while being backflushed with nitrogen at  20
ml/min (0.04 cu ft/hr).
  Dasorption and analysis. Desorber 1 (Fig. 3). The gas-cjjroma-

•Varian Aerogripn See. 1400
tFinnigan 1015C Quadrupole
tSysienu Industries 150
§M>IIipore Super-Q
 742   RESEARCH
                                      JOURNAL AWWA

-------
tographic oven was cooled to below 30C (86F) by leaving the
oven door open. The plug was removed from the desorber,
and the cap was removed from the trap; the trap was then in-
serted into the desorber and locked into place. The trap-back-
flush flow fitting was then locked  into place on the trap-exit
flow fitting and backflushed with nitrogen at 20 ml/min (0.04
cu ft/hr) for 3 min between  125C and 130C (257F and 266F).
The trap-backflush flow fitting was removed with the trap still
locked into place, the oven  lid was closed, and the oven was
rapidly heated to its normal or initial operating temperature. A
gas-chromatographic analysis was carried  out  under these
conditions.
  After analysis the trap was removed by: (1) inserting the
trap  vent into the trap-exit fitting  (to vent-inlet system), (2)
removing the trap, (3) resealing  GC-inlet system with the
plug, (4) removing the trap vent, and (5) resealing the trap in-
let with the cap.
  Desorber 2 (Fig.  4). The gas-chromatographic oven  was
cooled to below 30C (86 F).  The needle was inserted into the
liquid-inlet system  on the gas chromatograph. The trap was
then inserted into  the desorber and locked into place. The
trap-backilush flow fitting was locked into the trap-exit flow
fitting and backflushed with  nitrogen at 20 ml/min (0.04 cu ft/
hr) for 3 min between 125C and 130C (257F and 266F). After
desorption and  sample transfer  were completed, the needle
was removed from the liquid-inlet system, the oven lid was
closed, and the oven was rapidly heated to the initial operating
temperature. Gas-chromatographic analyses  were performed
under these conditions. After sample transfer the trap  was
removed and sealed for future use.

Investigation of Method  Parameters
  Initial studies were carried out to determine the  volume of
purge  gas needed  for quantitatively extracting selected
volatile  materials from a water sample. The purging device
was charged with 5.0 ml (0.3 cu in.) of an aqueous solution
containing methylene chloride, chloroform,  benzene, and 2-
butanone concentrations, each in excess of 10 mg/1 (10 ppm).
  As the solution was being purged with  nitrogen, 3-^1 ali-
quots were periodically withdrawn for analysis by direct-
aqueous injection. Analyses were performed on the aqueous
mixture until the concentrations of the dosed materials were
reduced to or below the limit of detection, approximately 100
Mg/1 (10~l ppm). This experiment  was initially performed
with a purge-gas-flow rate of 20-ml/min (0.04-cu ft/hr) nitro-
gen. The ilow rate was reduced 65 per cent to 13 ml/min (0.03
cu ft/hr), and the experiment repeated. The percentages of
the dosed compounds remaining in  the aqueous phase with
respect  to the purge volume are listed in Table 2.
  Those trap saturation volumes reported in Table 1,  desig-
nated by footnote t, were obtained by Bellar and Sigsby9 for a
dry-air system. To determine what effect, if any, water  that is
inherent to the system reported herein, would  have on the
saturation volumes, the authors redetermined the volumes
using water-saturated nitrogen as the purge gas; little if any.
change  was  observed.  The saturation volumes for several
organochlorme  compounds, not prsviously  reported, were
also determined under this condition.
  The purging-and-trapping system was tested with selected
industrial solvents over a  wide range of concentrations.
Ideally the response for each compound would be linear over
the entire concentration range. By using the standard solu-
tions and operating parameters previously described, the
autho'rs obtained the  data  listed  in Tables 3-6.  The peak

DECEMBER 1974
 height of each compound was measured and divided by the
 concentration to give  the slope between 0,0 and each data
 point collected. Response curves for four common organic
 solvents are shown in Fig. 5. The standard deviations from the
 mean slope are also listed in the tables.
   To determine the effect of variation in the physical proper-
 ties of individual compounds on the efficiency of the system,
 the authors tested a homologous series of n-alkanes. The test
 mixture consisted of n-C5 to n-C15 in organic-free water. This
 mixture was analyzed according to the prescribed procedure
 using a Tenax trap. Tenax was used as the adsorbent because
 it has a higher thermal stability than Chromosorb 101 and can
 be operated  at the temperatures required for desorbing the
 higher molecular-weight alkanes. To determine the purge
 volume  required for quantitative transfer of hydrocarbons
 over the wide boiling  range, successive fractions were col-
 lected at ambient temperature (19.5O [67F) and analyzed by
 flame ionization (FID) gas chromatography using an SE-30
 column (See Table 7). The test was  repeated at an elevated
 purging temperature (65C)  [149F] (Table 8).
   When the method was applied to a sample from a local sew-
 age plant which serves a diverse industrial area, the compli-
 cated FID  gas chromatogram shown in Fig. 6 resulted. The
 sample  was analyzed again using the microcoulometric detec-
 tor which gave the chromatogram shown in Fig. 7. The com-
 pounds identified  in  the chromatograms were confirmed by
 GC-MS.

 Results and Discussion
   The  data in Table  2  show that it  is possible to purge the
 water insoluble (<2 per cent soluble) compounds from 5 ml of
 water using < 150 ml (9 cu in.) of nitrogen. A decrease in the
 purge-gas flow rate of 65 per cent indicated that a slight in-
 crease in the volume of purge gas is  needed for quantitative
 transfer. Water-soluble materials whose partition coefficients
 do not favor the gaseous phase are  only qualitatively trans-
 ferred regardless of the purge volume.
   Trapping. Judging from the data reported by Bellar and
 Sigsby9 and other data exhibited in Table 1, one can see that
 organics contained in small volumes of water-saturated nitro-
 gen can be concentrated. It is apparent from these data that
 compounds with a retention index > 500 can be quantitatively
 purged and trapped. Retention indices given in the literature
 on porous polymers11'13 make it possible to predict trap satura-
 tion volumes for a wide variety of organic compounds. Since
 most hydrocarbons and substituted hydrocarbons commonly
 present in  wastewaters have retention indices >500, porous
 polymers were used in developing this method.
   Water has a retention index of < 300 and is not quan-
 titatively trapped by porous polymers. Therefore, gas-chro-
 matographic  columns and  detectors adversely  affected  by
 water can be used with a minimum of interference.
   The statistical data generated in Tables 3-6 reflect an ac-
 cumulation of errors for  the entire method. After one con-
 siders the  number of manipulations  involved and that gas-
 chromatographic errors are generally  ±3 per cent, it appears
 that this is, indeed, a useful method. Further study of these
 data indicates that the majority of the errors are caused by the
 volumetric-dilution procedure. The larger the pipet used to
 withdraw aliquots of the stock solution, the larger the error. A
 buret may  be a more suitable device for  delivering volatile
 solutes.
   For the compounds studied, based on data in tables 2-6, the
authors   estimated that  purging  transferred  at  least ()')

               T. A.  BELLAR AND J.  J. LICHTENBERG   743

-------
P -  cent of the volatile, water-insoluble compounds from the
aqueous phase to the gaseous phase. The data in Tables  3-6
a;id some unreported duplicate data show that the purging
etfiaency is identical from 2 500 ^g/1 (2 500 x 10~3 ppm) to at
kast  6 /Ag/1  (6 x 10~3 ppm).  Therefore, the  compounds
studied can be quantitatively determined over that concentra-
tion range.
  Further study of  the data in Table 7  indicates that  the
aikanes up  to €9 can  be quantitatively purged using < 500 ml
<30 cu in.) of purge gas. Purge volumes exceeding  1.5 1 (91.5
c- in.) failed to transfer 100 per cent of the Cn through  C}S
aikanes. Raising the  temperature of the purging device and
sample (Table 8) extended the useful range of the method to
C,, hydrocarbons. If a water sample contains volau'les over
the entire boiling range represented by these data, it may be
necessary to trap two fractions in order to perform a complete
quantitative analyses on the sample. This is apparent from the
data in Table 1 that show that compounds with a retention in-
dex of <  600 will saturate the trap and be vented before  the
high boiling materials are quantitatively purged.
  Swnp<« preservation. Because of the volatility of the organic
materials detected by this method, common sample-preserva-
tion techniques  are inadequate.3-'* The simplicity of the trap-
and-purging  device  makes it  possible,  however,  to collect,
purge, and trap the sample at the sampling site. The trap and
contents can then be sealed and shipped to the laboratory for
analysis, and thus, the need for sample preservation is elimi-
 nated.
   Application of  the method. Judging from the  experimental
data reported in  this article, one may see that this method has
great potential for the analysis of trace-volatile organics con-
 tained in a wide variety of water sources.  For quantitative
 determinations the method is limited to organic compounds
 that are  <  2  per cent soluble in water and boil below 200C
 (392F). Significant qualitative enhancement of compounds
 whose boiling points exceed 200C (392F) can be expected
 wnen the sample is  heated. The method is useful from 1 to
 2 500 jig/1 (10~3 to 2 500 x  10~3 ppm) with the use of most gas
 ctromatographs. At concentrations exceeding 2 500 /ig/1
 (2 500 x 10~3ppm) chromatograph-column flooding and non-
 linear-detector responses generally  occur. Since direct-
 aqueous injection techniques are useful down  to  1 000 /xg/1
 (\ 000 x 10~3 ppm) the two methods can be usedtogether to
 perform analyses over  a  wide  range of concentrations.  For
 water-soluble compounds the  distillation technique should
 provide  the  supplemental methbdology needed to analyze
 most industrial effluents and natural waters.
   \ wide variety of wastewater samples were analyzed using
 ti;-; described method. The chromatograms (Fig. 6, 7)  show
 the results of one such analysis.  Qualitative identifications
 ^ere  made using  desorber 2 and  a GC-MS system.* The
 quantitative analyses were obtained using desorber 1 with a
microcoulometric detector. Only one of the peaks in the FID
chromatogram have been  identified. At the sensitivity ranges
s.nown, only  the chlorobenzenes are likely to appear on  the
FI~' chromatogram.
  The method worked well except for the following: one sam-
ple collected from a sewage-treatment plant foamed ex-
 :assively and caused  water to be transported from the purging
Jev.c* into the trap. Decreasing the sample size from 5 to 3 ml
(0.3 to 0.2 cu in.) or using  the foam trap eliminated this  prob-
lem. Water entering the trap causes nonquantitative trapping
*Snni|*n System

744  RESEARCH
Reprinted .ir.i -orTl^ced as .1 part 3:
  Journal'American Wjcer Works Assn.
   Vol. 66 NJ.  12   Deceober [97i
         ?r'..-.:ii In U.S.A.
and severe gas-chromatographic interferences.
  When water samples contained gross amounts of water-sol-
uble organics, a sufficient quantity of these materials was col-
lected in the trap for detection. When only water-insoluble
materials were present in the sample, it was found that the
purged water could be withdrawn with a syringe and the purg-
ing device could be recharged for successive analyses. When
large concentrations of water-soluble organics were present, it
was necessary to dry the purging device in an oven at HOC
(230F) before an interference-free successive analysis could
be performed. Other researchers14-l5 have reported on similar
methods for the analysis of aqueous samples; their work has
been primarily qualitative.
  This current work has shown that the method can be used
for the quantitative measurement of a wide variety of water-
insoluble compounds whose boiling points are < 150C (302F).
By slightly modifying the method, one can also quantitatively
measure materials that boil at approximately 200C. Qualita-
tive sample concentration occurs for a wide variety of other
materials for which  quantitative measurements could possibly
be made if recovery factors were experimentally determined.
Vinyl chloride is one compound of considerable interest that
can be determined  by this method. Analytical conditions for
this specific application are under investigation.

Summary
  The method for  quantitative recovery and gas-chroraa-
tographic  determination of water-insoluble, volatile organic
compounds presented here provides a detection limit of ap-
proximately 0.5 Mg/1 for many compounds.

References
 1. National  Pollutant Discharge Elimination System, Proposed
    Forms and Guidelines for Information from Owners and Opera-
    tors of Point Sources, Pt. 2. Fed Rgtr.. 38:75:9783 (Apr. 19,1973).
 2. Ocean Dumping  Criteria, Pt 2. Fed. Rgtr., 38:94:12872 (May 16,
    1973).
 3. SUGAR, J.W.  & CONWAY, R.  A. Jour. WPCF, 40:9:1622 (Sep.
    1968).
 4. Tentative Recommended Practice for Measuring Volatile
    Organic Matter in Water by Aqueous-Injection Gas Chroma-
    tography, Annual Book of ASTM  Standards, Pt. 23,  Water.
    ASTM D 2908-70T, Atmospheric Analysis (1973).
 5. Methods for Organic Pesticides in Water and Wastewater. EPA,
    Natl. Envir. Res. Ctr., Cincinnati, Ohio (1971).
 6.  DUENBOSTEL, B.F. Method for Obtaining GC/MS Data of Volatile
    Organics  in Water Samples. Internal Rprt. EPA, Region II,
    Edison, N.J. (May 14, 1973).
 7. Procedure for Water Soluble Volatile Organic Solvents in
    Effluents and Streams. Org. Lab., Chern. Svces.  Br., Region 4,
    EPA. Athens, Ga. (Aug. 1973).                    '
 8.  Chlorinated  Organics and Hydrocarbons in Water by Vapor
    Phase  Partitioning and Gas Chromatographic Analysis. Method
    No. QA-466,  Dow Chemical,  Louisiana Div., Plaquemine,  La.
    (Jan. 1972).
 9.  BELLAR, T.A. & SIGSBY, J.E. Non-Cryogenic Trapping Techni-
    ques for Gas Chromatography, Internal Rprt.  EPA, Div. of
    Chem. and Phys., Research Triangle Pk., N.C. (1970).  '
10.  BELUAR, T.A. & SIGSBY, J.E. The Analysis of Light Aromatic Car-
    bonyls. Phenols, and Methyl Napthylenes in Automotive Emis-
    sions by Gas Chromatography, Internal Rprt.  EPA, Div. of
    Chem. and Phys., Research Triangle Pk., N.C. (1970).
11.  Chromosorb Century Ser. Bull., Johns-Manville, Celite Div.,
    Greenwood Pla2a, Denver, Colo. (Nov. 1970).
12.  Tenax-GC Bull. No. 24, Appl.  Sci. Lab., Inc. State College, Pa.
13.  HOLLIS, O.L. & HAYES, W.V. Jour. Gas Chromatog., 4:7:235 (Jul.
    1966).
14.  ZLATKIS, A. & LIEBICH. H.M. Profile of Volatile Metabolites in
    Human Urine. Clin. Chem.. 17:7:592 (Jul. 1971).
15.  NOVAK, J., ET AI_ Analysis of  Organic Constituents Present in
    Drinking Water. Jour. Chromatog, 76:1:45 (Feb.  1973).

                                       JOURNAL AWWA

-------
                         NO.  2
 CE COMPOUND TO CALIBRATE ION ABUNDANCE MEASUREMENTS
IN GAS CHRCMATCGRAPHY--MASS SPECTROMETRY SYSTEMS

-------
               Copyright 1976 by
Reprinted from ANALYTICAL CHKMISTItY. Vol. 47.1'ncn W-O. June 197.r>
)76 by the American Chemjcnl Society nn
 make empirical identifications of compounds in environ-
 mental, hiontcdical, nnd other type* of samples. Clenrly,
 correct identifications require some  consistency between
 reference spectra nnd observed spectra, and better quality
 abundance data would improve the effectiveness of all em-
 pirical search systems.

  1 Author to whom correspondence should h« addressed.
                                  In addition to an ion abun "ance calibrant, there is a need
                                for a reference compound to evaluate the overall perfor-
                                mance of a computerized GC/MS  system. \Ve have ob-
                                served spectra with acceptable  ion abundances but, be-
                                cause of poor resolution adjustment, broad peaks that were
                                interpreted by the data system as multiplets. A reference
                                procedure would allow an operator to validate the perfor-
                                mance of the  GC column, the sample enrichment device,
                                the ion source, the ion detection circuits, the analog-to-dig-
                                ital converter, the data reduction system, and the data out-
                                put system. The application of this procedure would en-
                                hance the overall quality of results emerging from the sys-
                                tems in use.
                                 There is a special need to closely monitor the perfor-
                                mance of the  RF quadrupole  mass spectrometer. Unlike
                                the magnetic deflection spectrometer, the active ion sepa-
                                rating device of the RF field spectrometer, the rods, is di-
                                rectly contaminated during operation. After prolonged op-
                                eration, the rods are subject to severely  degraded perfor-
                                mance which usually affects the region above 300 amu first.
                                Often this degraded performance is not detected because
                                there  is no generally accepted performance standard  to
                                form the basis for such judgments.
                                 The  Environmental  Protection Agency  has developed
                                and used experimentally a performance  evaluation/abun-
                               dance calibration procedure for the last several years. A set
                               of chemical and physical properties criteria for a  reference
                               material was developed and a number of likely candidates,
                               including PFK and PFTBA, were tested. The compound
                               decafluorotriphenylphosphine (DFTPP)  was  selected as
                               the one which met most of the criteria.
                                 This paper reports the criteria on which t.he compound
                               was selected, its mass spectrum, some physical and chemi-
                               cal properties, and some performance data that were col-
                               lected over the last few years. An RF field mass spectrome-
                               ter, which has been tuned  to give the suggested inn abun-
                               dances in the reference compound spectrum will,  in gener-
                               al, generate mass spectra of organic compounds which nre
                               very similar to spectra generated  by other types of mass
                               spectrometers. Thus RF Held mass spectra become directly
                               comparable to  spectra of compounds in collections which
                               have been obtained with otjier types of mass  spectrome-
                               ters.

                                               EXPERIMENTAL
                                 Material*. All chemicals unii solvents wore obtained from ofim-
                               mcrcinl suurcet. DefnlluorolriplicnylpluiNphiiiti woa prrjinrcd ac-
                                                        ANALYTICAL CHEMISTRY. VOL. 47. NO.  7, JUNE 1975 • 093

-------
   conlinq to the procedure of Dun (/); m.p. (<'i -M"; Anal. Cnlrd f(ir
   CiHHr.K,,,!': C, .1H.KO; II, 1.1.1. Kound: C. -trU)'., 49.0.1; H. 1.10. I 07.
   Purity, lu'xcd on flame iiiniy.nlinri detector t"H chruniHl> interfaced to lh« spectrometer by an all-
'   glass jet type cnricliment device and an all-i;l;is.s transfer line. Con-
   trol of the quudrupolc rod ma'ss .set voltages, data acquisition, data
   reduction, and data output was accomplished with a System in-
   dustries data system which employed a Digital Equipment Corpo-
   ration PDP-8/E mini-computer and a l.C million word Diablo disk
   drive.
     All of tl-c systems referred to in Table III also used this spec-
   trometer and data system which has a user option to integrate ion
   currents at one or more- (maximum of ten) 0.1-amu intervals be-
   tween each integer mass. The maximum ion current value is .select-
   ed for each amu by the control program and abundances of non-
   integer ionic masses are measured correctly. With DFTPP, this op-
   tion was not used since the ions in the spectrum of DFTPP have
   masses very close to  the integer values (e.g., M* = 441.997).
     The gas chromatograms and mass spectra were displayed on a
   Tektronix Model 4010 cathode ray tube or a Houston Instruments
   model DP-1 flatbed plotter.  .
     Gas Chroraatography. Most of the work reported in this paper
   was carried out using a 6-ft X 2-mm (i.d.) glass column  parked
   witK  1.95% QF-1 plus 1.5%  0V-17  on 80/100 mesh Supelcoport.
   The flow rate was about 30 ml/min; column temperature, 180°; in-
   jector temperature, 210°; and interface oven-transfer line tempera-
   ture, 200-210°. The compound decafluorotriphenylphosphine was
   also chromatographed on  a variety  of other columns of van-ing
   length and stationary phases. In general ihese were 4-8 ft, metal or
   glass,  100-250° column  temperatures, and 20-35  ml/min flow
   rates.  Stationary phases  were 3% SE30,  5.5% OVl. 3-5%  OV17,
   2-6% OV101, Dexil 300, and 0.1% OV210.
     Chromatography  was also successful on a 100-ft, 0.02-in. (i.d.)
   support coated open tubular column coated with QF-1. In general,
  • retention times of 4-10 minutes  were observed. Cross-linked po-
   rous polymer packed columns were not suitable for this compound.
   Similarly, a 7-ft coiled glai>s column  (i.d. 2 mm) packed with 10%
   free fatty acid  phase on 60/80 mesh chromosorb \V gave poor re-
   sults.
     Procedure. A stock solution of DFTPP at 1 mg/ml (1000 ppm)
   concentration in  acetone was prepared. This stock solution was
   shown, by repeated  analyses, to be 97%+ stable after six months.
   and indications are  it will  remain usable for several years. An ali-
   quot of the stock solution was'diluted to 10 up/ml (10 ppm) in ace-
   tone. The very small quantity of material present in this very di-
   lute solution is subject to depreciation because of adsorption on
   the walls of the glass container, reaction with trace impurities  in
   the acetone, etc. Therefore the dilute solution was used fur only a
   short term, i.e., 1-2 weeks.
     The gas chromatographic operating parameters were adjusted  to
   permit the acquisition of at least four complete mass spectra dur-
   ing the clution of the DKTPP. The mass/charge scale of the mass
   spectrometer was calibrated according to ihe standard procedure
   provided by the manufacturer. The muss spectrometer and data
   system were prepared for CC data acquisition using the following
   parameters: mass ran^c. 3.1-500 amu; electron energy,  70 eV; trap
   current. 250-500 »«A; preamplifier sensitivity, 10~7 A/volt; electron
   multiplier voltage,  3000  volts; and  muss spectrometer manifold
   temperature, 100°. Under these conditions, the ion source temper-
   ature of the Finnigan mass spectrometer  in not  known. The pres-
   sure in the spectrometer WHS about 1<>~' Torr nnd the base line was
   adjusted with the automatic zero program. The spectrometer dula
   system was set to integrate  the preamplifier signal for 8 msec  at
   each integer ma.is unit. Alternatively, the  iii(ej;nilion time as a
   function of si|;nnl strength option was iililixrd. This  will  \tf de-
   scribed in detail in a future publication (2).
     An injection of JO n;; CJ jj) of the dilute standard  wan mndc nnd
   datn Acquisition WHS In-guii iiltcr  most of I lie solvent WH.H pumped
   from (he spectrometer. Duiii 4ic(|in.silinn wiis rum-hided niter edi-
   tion of the DrTIT. The  m.iiw s|*clriiin of 1)1*1*1'!' was oliliiiiicd
   by KelcvtiiiK » r-pei-trniii nimilxT on the front side of the C>C, |»enk
   •3 nenr tin' upvx n.s possible. A lmrk|;rniiiul s|XTlruin WHS M-leclcd
   from ono of the spectra immediately (irecfilnn;  the OFTI'I' |H-uk.
   Severul uprrtra were sometimes plotted in un attempt  to find one
 which Hi the nlxindfinrc rrilcri.-i  If no upertMim could l»o oliliuned
 which fit  the criterin, the n«l nnd ion source potent i.ilt were nd-
 justed ns in the in;iniifnclurcr'» lune-up nriiwdiiri'. ll llu< failed to
 priKlucc Ihe correct  spectrum, more extensive mmnU-nance  «:IH
 performed. This was immlly cltiininu the ion source nnd/or the
 qundrupolc rods. These measures usually corrected the m.-ilcondi-
 lion and a spectrum of DFTl'l' could be obtained which fit the cri-
 teria.
            RESULTS AND  DISCUSSION
   The  results of several  recent studies illustrate the need
 for  a standard  relative abundance calibration procedure
 and pefonnance evaluation standard. A study was reported
 in 1973 (3) of calibration data from various types of mass.
 spectrometers. Relative abundance data were reported for
 an aliphatic hydrocarbon, n-hexadccane, and  an alkyiated
 aromatic hydrocarbon, 1-phcnyl undecanc. The participat-
 ing laboratories introduced these samples with convention-
 al batch inlet systems into a variety of single and double fo-
 cusing  magnetic  deflection  and  several  RK  quadrupole
 spectrometers. Selected data from  that study are  given in
 Table  I.  Measurements at  the selected ions agree reason-
 ably well below about mas* 100. Above mass 100, there is a
 clear indication of reduced sensitivity with the quadrupole
 spectrometers. This trend supports the widespread  idea
 that quadrupole spectrometers are significantly less sensi-
 tive than magnetic deflection spectrometers  at the  higher
 masses. The data above mass 100 obtained with the 21-491
 and MS-902 spectrometers reveal the well known fact that
 magnetic deflection  spectrometers  are  susceptible  to re-
 duced  high mass sensitivity also. This may be due to em-
 phasis  on low mass sensitivity during ion source tuning or
 performance degradation due to contamination of the ion
 source.
   In late 1972, samples  of DFTPP  were sent by us  to a
 number  of EPA and other laboratories. This survey was
 conducted to  obtain relative abundance comparisons up to
' mass 450. In addition, it  was requested that the sample be
 introduced with a GC inlet system and any GC column that
 was convenient for the  participating laboratory.  The re-
 sults from magnetic deflection systems are shown in Table
 II and  from RF quadrupole systems in Table III. The ions
 selected  for  comparison  are spaced at approximately 75
 amu intervals up to mass 275 and include, in addition, the
 molecular ion (M*) at mass 442 and the molecular ion con-
 taining a single 13C atom at mass 443. The theoretical 443/
 442 percentage is 19.8%.
   Relative abundance data for DFTPP from three  of the
 magnetic sector instruments is in very good agreement and
 all four magnetic instruments  produced acceptable  values
 for the (M+ + 1)/M+ percentage. The relative abundance
 data from the 21-490 may be an example of ion source tun-
 ing  to  emphasize the molecular ion region  or perhaps it
 merely reflects the selection  of a spectrum  number  too
 close to the front of the peak. In the latter event,  the  mo-
 lecular ion would have been observed after the concentra-
 tion of the DFTPP in the ion source had increased signifi-
 cantly.
   The  relative abundance  data for DFTPP  from the  RF
 quadrupole spectrometers were much less consistent. Lab-
 oratory No. 1  reported the base peak as mass 51, laborato-
 ries 2-7 reported  the base  peak as mass  198, laboratory 8
 reported the molecular ion ns the base peak, and laborato-
 ries 9-11 found muss 09 (CK.!*) n.s the base ponk. The range
 of abundance measurements at nny  RI'VIMI muss was pencr-
 nlly mudi larger with  I lip  RF qundrupolc spectrometers.
 For oxiintple,  the three magnetic deflection spectrometers
 that nu'astircd mass 198 us  I ho  base peak had n range of 21
 relative abundance units at mass 51. The six quadrupolcs
 that measured mass 198 as the  base penk had a range of 39
   •98 • ANALYTCAL CHEMISTRY. VOL. 47.  NO. 7, JUNE 1975

-------
 Table I. Selected Relative Abundance Dulu for llexudccanc Measured with a Variety of Mass Spectrometers'1
                                                        Kclatlv« ibuuJanc*,*

         M«u
          57
          71
          85
          99
         113
         127
         141
         155
         226
   • Dala taken from Reference 3. * Sinplc focusing sector magnetic deflection spectrometer. r Double focusing (electrostatic ami magnetic
 fields) modified Nier-Johnson spectrometer. '•Double focusing Maliauch-Herzog geometry spectrometer. 'Radio frequency t|ii;idrupole
 spectrometer.
CH-7*
100
60
37
12
7
5
4
4
8
HMU-6»
100
55
40
12
8
7
5
5
12
21-4-W*
100
73
40
14
9
7
G
5
11
21-49K
100
65
45
12
7
5
3
3
3
21-4'jf
100
75
52
16
9
6
6
6
11
21 -non'
100
72
48
11
7
6
5
5
9
MS-DOS*
100
66
37
10
6
4
2
2
3
1015'
100
54
32
7
3
2
1
1
1
101 5"
100
60
35
8
4
2
1
1
1
 Table II. Selected Relative Abundance Data for DFTPP Measured with Single
     Focusing Magnetic Deflection Spectrometers and GC Inlet Systems
                                                          Percent relative abuadaacc at majs-

Sp«ctToro«l*r
Varian CH-7
Varian CH-5
Nuclide 1290G
DuPont 21-490

51
40
60
34
12

127
42
52
37
13

198
100
100
100
34

275
26
24
29
11

442
92
95
86
100

•U3
20
19
17
21
2
443/442' 10
21.7
20.0
19.8
21.0

 Table III. Selected Relative Abundance Data for DFTPP Measured with
     Fianigan 1015 RF Quadrupole Spectrometers and GC Inlet Systems
                                                    Percent relative abundance at ma«
Lab
1
2
3
4
5
6
7
8
9
10
11.
SI
100
81
53
53
92
86
57
14
66
93
97
127
49
50
68
48
55
40
43
19
80
57
85
193
98
100
100
100
100
100
100
42
76
85
65
275
20
13
24 .
19
22
28
16
13
19
11
11
442
51
33
31
64
57
56
48
100
47
20
2.5
443
9
7.5
5.5
12
12
10
10
91
13
4
2.5
443/442 * 102
17.6
22.7
17.7
18.8
21.0
17.9
20.8
91.0
27.7 .
20.0
100.0
 units at mass 51. All four magnetic deflection spectrome-
 ters produced  molecular ion  measurements between  86-
 100%; the quadrupolc values for the molecular ion ranged
 between 2.f>-100%.  The  values of the (M* +  1)/M+ per-
 centage  from the four magnetic  deflection spectrometers
 had a standard deviation ot'O.S"... The same values from the
 quadrupole spectrometers had a standard deviation of 3%
 after rejection of the 91% and 100% observations.
  The more diffuse nature of the RF quadrupole  abun-
 dance measurements was probably due to a variety of caus-
 es including the presence of generally less experienced  op-
 erators, the failure  of some operators to utilize ion  abun-
 dance  calibration  procedures, inadequate  ion  source  or
 quadrupolc  rod maintenance, more difficult qundrupole
 tune-up adjuMments, and the selectipn of spectrum num-
 bers too close to the front or apex of a (1C peak.
  The hcxadecane spectrum was measured in this labora-
 tory with on RF iniaclrnpolc niter the spectrometer was ad-
justed to give a DFTPP spectrum similar to that produced
by I he Vurinn and Nuclide magnetic deflect inn spectrome-
ters. The musses nnd relative abundances that correspond
to those in Table I were: 57. 100; 71, U5; 85, ;!!); Oil,  Kl; 113
            1J-_>lJ, 10.
   \Ve concluded  that the RF quadrupole spectrometer
 could be maintained, without unreasonable effort, in a con-
 dition that would produce mass spectrometric fragmenta-
 tion patterns that were very similar to patterns produced
 by other types of spectrometers. However, it was also clear
 that a standard relative abundance calibration procedure
 and performance  evaluation  standard was  required  and
 that use of this standard would benefit users of magnetic
 deflection spectrometers also.
   Criteria for the Ideal Reference Compound. The
 ideal  reference compound should possess a number of im-
 portant properties. It should be available in very pure form
 as a crystalline solid. This is necessary to facilitate accurate
 weighing and the- preparation of standard solutions to eval-
 uate GC/MS system sensitivity  in terms of signal to noise
 for a  given quantity. The compound should  have high ki-
 netic and thermixlynamic stability nnd be soluble in a vari-
ety of common organic solvents to facilitate gus chromatog-
 rnphy. The material should be very easy to gas  chrumato-
grapb on a wide variety of columns of differing  polarity.
This property would encourage its application on whatever
column was of particular importance in a given laboratory.
The muss  spectrum of the compound must display  an
                                                           ANALYTICAL CUCMISTRY. VOL. 47. NO.  7. JUNE  1975  •  997

-------
 ebundnnt molcculnr or fragment i«n ncnr mnss .r>00. This is
'an extremely important factor since  mnny comjH>unds of
 environmental nnd biomedical significance hiivc ions in the
 400-500 umu range.  The ion must be very abundant in
                      i puKoat
                                        og

                                          9
                                          I
                                          8
                                          8
                                          D
                                          $5
                                          $
                                          .8
                                         12
                                          8
                                          8

                                        - .8
                                        , .8
u.
Q
w

I
                                          .P
                                          .8
                                           a
                                            -  2
                                               •
order to cnsily evaluate the system acnsitivitv and resolu-
tion in the high muss region. The commonly use<) mass cali-
bration compounds PFK and IM-'THA have ions HI this re-
gion, but these nre very inadequate because of their very
low relative abundance. For example, a 100% reduction in
spectrometer  sensitivity at mass  500 is reflected in  the
spectrum of PFTBA by a change in relative abundance of
the mass 502 ion from 2% to 1%. In contrast, the spectrum
of the  reference standard should not be dominated by a
single very abundant ion which tends to saturate the detec-
tor and reduce all other ions to very small relative abun-
dances.  Most fluorinated  aliphatic  compounds,  e.g.,
PFTBA and PFK constituents, suffer from the dominance
of the  mass 69 CV-.\+ ion. An even distribution of ions of
even relative abundances over a wide mass rnn^e is most
desirable. On  the  other hand,  the compound should  not
possess too many ions which might cloud a spectrum with
too much information to allow a fast evaluation of the sys-
tem performance.
   It was clear that the fluorinated aliphatic compounds in-
cluding PFK and PFTBA were not suitable because of sev-
eral serious limitations. n-Hi-- tdecane is widely used as  the
standard of reference in hydrocarbon type  analyses in  the
petroleum industry, but its low molecular weight.  226, and
the generally low relative abundance of the molecular ions
of aliphatic hydrocarbons rule out this type of ?tandard.
Cholesterol is  a crystalline compound of reasonable molec-
ular weight but it is difficult to chromatograph  without
derivatization. Methyl stearate is often used as a test com-
pound  but  it is unacceptable  because of its molecular
weight, 298, and the low relative abundance of the molecu-
lar ion. Perfluorodecalin was recently proposed  (•»') as a
mass calibration standard for low resolution mass spectra.
One of its attributes is that the relative contribution of
mass 69 to the total ionization is much less than for other
fluorinated aliphatics. Nevertheless, the compound is a vol-
atile liquid with no ion of greater than 10°o relative abun-
dance above mass 293. Perfluoroalkyl-s-triazines and relat-
ed compounds (5)  have been used to excellent advantage as
very high mass calibration standards for the mass to charge
scale.  They suffer similar disadvantages of dominance by
mass 69 and large gaps where no abundant  ion is observed.
Triphenylnaphthalene (6) was reported as a useful mass to
charge scale  calibrant. This compound has a molecular
weight of 356 and produces a large number of ions includ-
ing several abundant clusters.
  " The compound bis(perfluorophenyl)phenylphosphine 1
(or decafluorotriphenylphosphine. DFTPP) was one of a
number of compounds  evaluated as a possible iojn abun-
dance  calibration reference compound and standard  for
performance measurements. Its spectrum is shown in Fig-
ure 1. The compound  meets nearly all of the criteria  de-
scribed previously. Its spectrum contains  relatively abun-
dant ions at about 75-ainu intervals (Tables II and III)  be-
tween masses 51 and 275. It is deficient in  that there is no
abundant ion in its spectrum between mass 275 and 442.
The molecular ion at mass  4-12 is very abundant  but does
not dominate, mid  there are not too many ions that would
preclude rapid inspection and evaluation of a spectrometer
performance.
                 no*
           Ai VTW*»«I <*uca«CTOV \/ni  tt  MA  7
                                                    1O7S

-------
     Proposed Compositions of Ions in I lie Spectrum of
   DFTPP. Kx.id inns*; niiMVirr.-in<»iil.s nn.ri i-nrrcsprinds to loss of a phenyl group.
   The ion ;it mass 27~> has the composition (C,;H.-,)(CfiFfi)l>'1'
   which results from Id-.-  loss of a single  perfluoropbenyl
   group from the molcr;i!;..- ion. We  propose the fragmenta-
   tion process in whirh  this ion either loses its phenyl group
   to form the mass I9S ion, or loses hydrogen fluoride to form
   the tetrafliioroph(is|iha?')li: ion of mass 2f>.'j.
    Mass  127 is perhaps  the phenylfluorophosphine  ion
  CgHsP+F. The ions of m.vses .77, 69, and 51 are well estab-
  lished as the phenyl, CF;+, and C4H;t+ ions. The latter is a
  decomposition  product of the phenyl ion and the CFn4 ion
  is produced  by extensive rearrangement of a perfluoro-
  phenyl ion.
    Relative Ion Abundance Criteria. It was our goal to
  arrive at  a set of relative abundances for  DFTPP that
  would be a standard for  performance evaluations and a
  guide for  ion abundance calibration. The data collected in
  the 1972 survey (Tablf- II and III)  as well as hundreds of
  repeated measurements in this and several other EPA labo-
  ratories were the basis for  these criteria. It must be empha-
"  sized that the data from the 1972 survey were taken direct-
  ly from the computer program generated plots  or digital
  printed data when available  and that  the criteria are  in-
  tended to apply to the same output. The data handling sys-
  tem  of a modern GC/VuS is an integrated part of the total
  system, and the data system performance must be included
  in the overall evaluation. Clearly, the computer generated
 output is the  most convenient for the operator to use in the
 evaluation.
   The majority of measurements found  mass 198 as the
 base peak and this  vv.ns  selected as the basic criterion
 (Table IV). All other criteria were  developed using  only
 those spectra which h;\d mass 198 ns the base peak. Abun-
 dant ions  were  located at approximately 75 amu intervals
 above ami below mass 1!>S. These wore masses 51, 127, and
 275 nnd they were included in  the criteria to provide a men-
 sure  of system  sensitivity  ut  regular intervals  throughout
 the mass range.  The molecular ion at mass -1-12 and the very
 scarce ion  ;>t mass 36~> were selected  for the same purpose.
Abundant  ions at masses (!!). 77, 110, and 2f».r> were not used
 because the selected ions adequately measure system sensi-
tivity. Ma»s (?S» wns specifically excluded from the criteria
because its abundance frequently depends on  background
conditions thai  result from tin? use of J'KK, PFTHA, etc..
for mnsx/charge  scale calibrations.
   In spectra (Tables 11 and 111) thnt had muss 198 ns tin;
bu«c peak, .seven of the nine molecular  ion measurements
Table IV. Reference Compound Key Ions und
    ion Abundance Criteria
                        lou itiundjare criteria
              30-60% of m:iss 198
              Less than 2'.V. of mass 09
              Less than 2% of mass C9
              40-60% of mass 198
              Less thaji !'£ of mass 198
              Base peak, 100'',', relative abundance
              5-9% of mass 198
              10-30% of mass 198
              l%of mass 198
              Less than mass 443
              Greater than 40% of mass 198
              17-23% of mass 442
         51
         08
         70
        127
        197
        198
        199
        275
        365
        441
        442
        443
 were greater than 40% relative abundance. Therefore, this
 was selected as a reasonable lower- limit for the molecular
 ion abundance.  No upper 'iir.it was set. All r.ine spectra
 showed an ion of 1-3% at mass 365 and a system with ade-
 quate high mass sensitivity should detect at least a 1% ion
 at this mass. The average abundance for  mass 275 in  the
 nine measurements was 22% with a standard deviation (a)
 of 5%. This was rather low dispersion for a set of relative
 abundance measurements and suggests that the abundance
 at mass 275 is closely related to the arbitrarily constant  ion
 abundance at mass 198. This is consistent with the compo-
 sition assignments discussed previously. However, a toler-
 ance at mass 275 of ±5% was considered too small for rou-
 tine GC/MS applications. A criterion at mass 275 of 20 ±
 10% was selected by rounding the average relative abun-
 dance to the nearest ten percent and allowing a deviation of
 2.  Rounding
off and using a 2 
-------
In the mid nntl IMW mass rondos, similar resold! ion checks
were developed  i:\inj;  ions ronl.'iiijinj; a single  i:'C  ion at
massed 100 ami 70. In each insl.-mc-u, the ions an  very likely
fissioned the correct composition nnd the  thmrrticnl  per-
centages mny l»c compared with ihc experimental. At mass
190, nine mcaMi.-c:ncnis gave nn average of 8.(X\, nnd a —
2.3. The criterion suggested is 7 ± 2% which compares with
the tKeorcticnl  value of 6.6%. At mass 70,  ihr  iheoretical
value is 1.1%, hut most of the nine measurements nave near
zero values  for this ion.  Perhaps this wns caused l)y very
slight changes in  the base-line (threshold).adjustments. It
is very  difficult  to make accurate and precise measure-
ments of relatively non-abundant ions when observing very
small amounts (=s20  ng) in  fast (3—4 sec) spectrometer
scans. Therefore, for mass 70, we suggest a nominal criteri-
on of less than 2% of mass 69. This is mainly a check on ex-
cessive broadness or poor peak shape in the low mass re-
gion for those data systems that interpret broadness as ion
abundance.
  Because of the probable compositions of the mass  198
(CGF5P+) and mass 69 (CF.-1+) ions, it is unlikely that mass
197 and  mass C8 ions would  he  present. Indeec'  repeated
measurements  have shown  that  they  are -not present.
Therefore, we suggest that mass 197 should  be less than 1%
of the base peak, and mass 68 less than 2% of mass 69. Both
criteria are checks on excessive broadness and skew as dis-
cussed above.

                    CONCLUSION
  The set  of  relative  abundance ranges  proposed  for
DFTPP has been very  useful in evaluating the performance
of n number of (JC/MS systems. Those ranees nrc the basis
for the proposed standard ion abundance calibration and
provide a reasonable basis for comparing the output from
the wide variety of systems in use.
                ACKNO WLKDC MENT
  We express our sincere  appreciation to the individuals
and  laboratories that participated in the interl.iborntory
study. These included J. Peterson, Fish and Wildlife Ser-
vice; E. M. Chait, K. I. DuPont de Nemours &. Company; J.
C. Cook, University of Illinois; J. B. Knight, Finnifjan Cor-
poration; and F. Biros, J. Blazevich, H. Boyle, M. Carter, P.
Clifford, F. Farrell,  G. Muth, H. Rodriguez, D. C. Shew,
and A. Wilson of the Environmental Protection Agency.
                LITERATURE CITED

(1) S. S. Qua, R. C. Edmondson. and H. Gilman. J. Organome'-ai. Oem.. 24,
   703(1970).
(2) J. VV. EicTielberger. L. E. Harris, ana W L. Budde. to b« y^: >h»3: pre-
   sented at th« 22nJ Annual Conference en Mass Spectrome:r, and Aflied
   Topics, Philadelphia, PA. May 19-24. 1374
(3) American Society for Testing and Materials Committee 0-2. 21st Annual
   Conference on Mass Spectrometry and Al:,ed Topics. San Francisco.
   CA, May 20-25. 1973.
(4) B. S. Middleditch. Anal. Chem.. 41. 2092 (1969).
(5) R. H. Wallick. G. L-. Pecle, and J. B. Hyres. Anal. Chem.. 41.382 (1969).
(6) O. M. Schoengold and W. H. Stewart. Anal. Chem.. 44, 834 (1S72).

RECEIVED for review October 29, 1974. Accepted January
24,1975.

-------
                         CE NO.  3
MEASURING VOLATILE ORGANIC MATTER IN WATER BY AQUEOUS-INJECTION
                   GAS CHRCMATOGRAPHY

-------
         Designation: D 2908 - 74
         Standard Recommended Practice for
         MEASURING  VOLATILE  ORGANIC  MATTER  IN
         WATER  BY AQUEOUS-INJECTION  GAS
         CHROMATOGRAPHY1
This Standard it issutd under the Hied designation D 290S: the number immediately following (he designation mdicatesihc
tear of original adoption or. in the case of revision, the >ear of last revision. A number in parentheses indicates the vear of
last reapproval.
1. Scope
   1.1 This recommendeo practice covers the
general  considerations for the qualitative and
quantitative determination of volatile organic
constituents in water by gas-liquid chromatog-
   1.2  Direct aqueous injection  of samples is-
feasible at organic concentrations greater than
1  me/liter.  The  applicability of  the  method
can be extended to waters of lesser concentra-
tions by  evaporative techniques,  freeze-out.
solvent extraction, or carbon adsorption.4

2. Significance
   2.1  The major organic constituents in  in-
dustrial waste water need to be identified for
support of effective in-plant or pollution con-
trol  programs.  Currently the most practical
means for tentatively identifying and measur-
ing a  range  of  volatile organic compounds is
gas-liquid chromatography.  Positive identifi-
cation requires  supplemental  testing (for
example, multiple columns,  speciality detec-
tors. spectroscopv, or a combination of these
techniques).
3. Summary of  Method
   3.1  This recommended practice defines  the
applicability  of various  columns  and  condi-
tions for  the separation of paturally occurring
or synthetic organic:, or both,  in  an aqueous
medium for subsequent detection with a flame
ionization detector. After  vaporization,  the
aqueous sample is carried through the column
by an inert  earner gas.  The sample  compo-
nents  are partitioned between the carrier  gas
and a stationary liquid phase on an inert solid
support. The column effluent is burned in an
air - hydrogen  flame. The ions released from
combustion of the organic components induce
an increase in  standing current which is meas-
ured. Although this method is written for hy-
drogen  flame  detection,  the  basic technology
is applicable to other detectors if water does
not interfere.
  3.2 The elution times are characteristic of
th'e various organic components present in the
sample, while  the peak areas are proportional
to the quantities of the components. A discus-
sion of gas  chromatography is presented in
ASTM  Recommended"  Practice  E 260. Gen-
eral Gas Chromatography Procedures.6

4. Definitions
  4.1 The  following terms in  this recom-
mended practice are defined  in accordance
with ASTM Definitions D 1129, Terms  Relat-
ing to Water*:
  4.1.1 "ghosting"  or  memory peaks—an
interference,  showing  as a peak,  which ap-
   " This recommended practice is ur»dcr the jurisdiction of
ASTM Committee f>l9on Water.
  Current edition approved June 27. 1974.  Published July
1974. Originall* published as D :908 - 70 T. Last preview
edition D~:90S'- 70 T.
   B> publication of this standard no position is taken »'ih
rc^pe.n to  the xjlidit) of an\  patent rights m connection
!T)ere»ith. and the ^mencan Society for Testing and Mate-
rials doe> not undertake to insure anyone utilizing the it-">-
djrd jyjin^i liability for infringement of any Letters Patent
nur j—ume jnv  >uch liability.
   • The boldface numbers in parentheses refer to the refer-
ences appended to inn recommended onnicc.
   Refer aNo (o ASTM  Meihod D 24&0, Tesi for  Phenol*
m U jter b> Gji-Liyjid Chromotography. which appears'»
this publication
   ' Kor information on i«o of ihcte  concentration teen*
niuues. refer 10  ASTM Meihod D 2778. Solvent Extract**
01' Organic Miner from Water and ASTM Recommended
Practice D XIJ. Removal of Organic  Matter from Wiier
K  V.l.'jlcJ  ( jrrs.n  Adsorption, both of »hich appear i»
the lt-4 Annual Bouk of ASTM Standards. Pan 31.
   • 1V'4 Annual Hoot of ASIH Standard!. Part 42.
   ' 1974 Annual Soak of ASTU Slaadanis. Part 31.
                                            480

-------
                                              D 2908
 pears at the same elution time as the organic
 component of previous analysis.
   4.1.2 internal standard—a  compound  of
 known behavior added to a sample to facili-
 tate the analyses.
   4.1.3 noise—an extraneous electronic sig-
 nal which affects baseline stability.
   4.1.4 retention time—the lime that elapses
 from the introduction of the sample until the
 peak maximum is reached.
   4.1.5 relatice retention ratio—the retention
 time of the  unknown  component divided by
 the retention time of the internal standard.
   4.2 For definitions of oiher terms used  in
 these methods, refer to ASTM Recommended
 Practice E 355, Gas Chromatography Terms
 and Relationships/

 5. Interference
   5.1 Paniculate Matter— Paniculate or sus-
 pended matter should be removed by centrifu-
 gation or  membrane filtration if components
 of interest are not altered. This pretreatment
 will prevent both plugging of syringes  and
 formation of  condensation nuclei. Acidifica-
 tion will often facilitate the dissolving of par-
 ticulate matter, but the operator must deter-
 mine that pH  adjustment does not alter the
 components to be determined.
   5J Identical Retention Times—With  any
 given column and operating conditions one or
 more components  may  elute at  identical re-
 tention times. Thus a  chromatographic peak
 is only presumptive evidence of a single com-
 ponent. Confirmation requires analyses with
 other columns  with  varying physical  and
 chemical properties or spectromethc confir-
 mation of the isolated peak or both.
   5.3 Acidification—Detection  of certain
 groups of components will be enhanced if the
 sample is made neutral or slightly acidic. This
 may minimize the formation of nonvolatile
 salts in cases  such as the analysis of volatile
organic acids  and  bases and certain  chloro-
 phenols.
  5.4 Ghosting—Ghosting is evidenced by an
 interference peak that occurs at the same time
 ts that for a component from a previous anal-
ysis but usually with less intensity. Ghosting
occurs because of organic holdup in the injec-
tion  port.  Repeated water washing  with 5-»l
injections  between sample  runs  will usually
 eliminate ghosting problems. The baseline is
 checked at maximum sensitivity to assure that
 the interference has been eliminated. In addi-
 tion to water inieciions, increasing the injec-
 tion port temperature for a period of time'will
 often  facilitate the elimination of ghosting
 problems.
   5.4.1  Delated  Elution—Highly  polar  or
 high  boiling components may unpredictably
 elute several chrornatograms later and  there-
 fore act as an interference. This is particularly
 true with complex industrial waste samples. A
 combination  of repeated water injections and
 elevated column temperature  will eliminate
 this problem. Back flush valves should be used
 if this problem is encountered often. Carrier
 gas wetted  by  steam can  be used  to  reduce
 component holdup  in  some cases: however.
 column life may be seriously shortened. Pass-
 ing the carrier  gas through a pre-column con-
 taining  copper sulfate  (CuSO4 • 5H;0)  for
 wetting may  have a  lesser  effect  on substrate
 stripping (1).

 6, Apparatus
   6.1 Gas System:
   6.1.1  Gas  Regulators— High quality pres-
 sure regulators should be  used  to ensure a
 steady flow of gas to the instrument. If tem-
 perature programming  is  used,  differential
 flow controllers should be installed in the car-
 rier gas line  to prevent a decrease in flow  as
 the pressure drop across the column increases
 due to  the  increasing temperature. An un-
 steady flow will create an unstable baseline.
   6.1.2  Gas  Transport Tubing—New  rubing
 should be washed with  a detergent  solution.
 rinsed with cold water, and solvent rinsed  to
 remove residual organic preservatives or lu-
 bricants. Ether is an effective solvent. The
 tubing is then dried by flushing with nitrogen.
   6.1.3  Gas  Leaks—The gas system  should
 be pressure checked  daily for leaks. To check
 for leaks, shut off the detector and pressurize
 the gas system to approximately  103 kPa (15
 psi) above the normal operating pressure. Then
shut off the tank valve and observe the level of
 the pressure eauge. If the preset pressure holds
 for 10 min, the system can be considered leak-
 free. If the pressure drops, a leak  is indicated
and should be located and  eliminated before
proceeding further. A soap solution may be
                                        481

-------
                                             D 2908
used for determining the source of leaks, but
care must be  exercised to avoid getting the
solution inside the tubing or instrument since
it will cause a long lasting, serious  source of
interference.  Leaks may also occur between
the instrument gas inlet valve and flame tip.
This may  be checked by removing the flame
tip, replacing  it  with a closed fitting and re-
checking for  pressure stability  as previously
noted.
  6.1.4  Gas Flow— The gas flow can be deter-
mined with a bubble flow  meter. A  micro-ro-
tameter in the gas inlet line is also helpful.  It
should be recalibrated after each readjustment
of the gas operating pressure.
  6.2 Injection   Port—The  injection port
usually is insulated from the chromatographic
oven and equipped with a separate heater that
will  maintain  a  constant temperature. The
temperature of  the injection port should  be
adjusted  to  approximately 50  C above the
highest  boiling sample component.  This will
help minimize the elution time, as well as
reduce peak tailing. Should thermal decompo-
sition of components be a problem,  the injec-
tion port temperature should be reduced ap-
propriately.  Cleanliness of the  injection port
in some cases can be maintained at a tolerable
level by  periodically raising the temperature
25 C above the normal operating level.  Use of
disposable glass  inserts or periodic cleaning
with chromic acid can be practiced with some
designs. When using samples larger than 5 ul.
blowback into the carrier gas supply should be
prevented through use of a preheated capillary
or other special design. When using 3.175-mm
(0.125-in.) columns, samples larger  than 5 n\
may extinguish  the flame  depending  on col-
umn length,  carrier gas flow,  and injection
temperature.
  6.2.1 Septum—Organics  eluting  from  the
septum in the  injection port have been found
to be a  source of an unsteady baseline when
operating  at high sensitivity. Septa should be
preconditioned. Insertion ot" a new septum in
the injection  port at the end of the day for
healing overnight will usually eliminate these
residuals. A separate oven operating at a tem-
perature similar to that of the  injection port
can  also be used to process the septa. The
septa should be changed at least once a  day to
minimize  gas leaks and  sample  blowback.
Septa with TFE-fluorocarbon backings mini-
mize organic bleeding and can be used safely
for longer periods.
  6.2.2 On-Column Injection—While injec-
tion into the heated chamber for flash vapori-
zation  is the most  common injection set-up.
some  analyses  (for  example, organic acids)
are better performed with on-column injection
to reduce ghosting and peak tailing and to
prevent decomposition of thermally degrada-
ble compounds. This capability should be built
into the injection system. When using on-col-
umn injection a shorter column life may occur
due to solid build up in the  injection end of
the column.
  6.3  Column   Ocen—The  column  ovens
usually are insulated separately from  the
injection port and  the detector.  The oven
should be equipped with a proportional heal
and a  squirrel-cage blower to assure maxi-
mum  temperature  reproducibility and uni-
formity throughout  the oven. Reproducibility
of oven temperature should be within 0.5 C.
  6.3.1  Temperature   Programming—Tem-
perature  programming is desirable when the
analysis  involves the resolution of organics
with  widely varying boiling points. The  col-
umn oven should be equipped with tempera-
ture programming between 50 and 350 C with
selectability  of several programming rates
between  1 and 60 deg/min  provided. The ac-
tual column temperature will lag  somewhat
behind the oven temperature at  the faster
programming rates.  Baseline drift will often
occur  because of increased  higher tempera-
tures  experienced during temperature pro-
gramming. This depends on the  stability of
the substrate  and  operating  temperature
range.  Temperatures that approach the maxi-
mum limit of the liquid phase limit the oper-
ating range. Utilization of dual matching  col-
umns and a differential electrometer can min-
imize the effect  of drift: however, the drift is
reproducible and does not interfere with the
analysis in most cases.
  6.4  Detector—The combination  of high
sensitivity and a wide linear range  makes the
flame  ionization detector  (FID)  the usual
choice in trace  aqueous analysis. The flame
ionization detector is relatively insensitive to*
water  vapor  and  to  moderate  temperature
changes if other  operating parameters remain
unchanged. If temperature  programming is
used, the detector should be isolated from the
                                          482

-------
                                        flb
D 2908
 oven and heated separately to ensure uniform
 detector temperature. The detector tempera-
 ture should be set near the upper limit of the
 programmed temperature  to prevent conden-
 sation.  The detector should also  be shielded
 from air currents which could affect the burn-
 ing  characteristics of the flame.  Sporadic
 spiking in the baseline indicates detector  con-
 tamination: cleaning, preferably with diluted
 hydrochloric acid (HC1. 5 4- 95), and an ultra-
 sonic wash with  water is necessary. Chromic
 acid also can be  used if extreme care is taken
 to keep exposure times  short and if followed
 by thorough  rinsing. Baseline noise may  also
 be caused by dirty or corroded electrical  con-
 tacts at switches due to high impedance feed-
 back.
   6.5 Recorder—A 1-mV, l-s, full-scale re-
 sponse, strip-chart recorder, is recommended
 to obtain a permanent chromatogram. Chart
 speeds  should  be adjustable between  15  and
 90 in./h.
  6.6 Power Supply—A.  105  to  125-V,  a-c
 source of 60-Hz frequency suppling 20-A serv-
 ice is required as a main power supply for
 most gas  chromatographic  systems. If voltage
 fluctuations affect baseline stability,  a voltage
 regulating transformer  may be  required in
 addition to the one incorporated within  the
 chromatographic instrument.

7. Retgents and Materials
  7.1 Purity  of  Reagents—Reagent   grade
chemicals shall be used in all instances  for gas
purification,  sample stabilization, and other
 applications.  Unless otherwise indicated, it is
intended that all reagents shall conform to the
specifications of the Committee on Analuical
 Reagents of the American  Chemical Society.
where such specifications are available.' Other
grades may be  used, provided it is first  ascer-
tained that the reagent is of sufficiently high
purity to  permit  its  use without lessening the
accuracy of the determination.
  7.1.2  All chemicals used for internal stand-
ards shall be of highest known punty.
  7.2 Purity  of Water—Unless otherwise in-
dicated,  references to  water shall be  under-
stood to mean reagent water conforming to
Type I  of ASTM Specifications  D  1193.  for
Reagent Water.'
  7.3 Carrier Gas System—Only gases of the
highest  purity obtainable should be used in a
 chromatographic system designated for trace-
 organic  monitoring  in  water.  The  common
 carrier gases used with a flame ionization de-
 tector (FID) are helium  and nitrogen. Trace
 contaminants in even the highest purity gases
 can  often affect  baseline stability and  intro-
 duce noise. Absorption columns of molecular
 sieves (14 by 30-mesh) and anhydrous calcium
 sulfatc (CaSO4, 8 mesh) in series between the
 gas supply tank and the instrument will mini-
 mize the effect of trace impurities. These pre-
 conditioning  columns, to remain effective,
 must be cleaned by back flushing them with a
 clean gas (nitrogen,  helium) at approximately
 200  C. or they must be replaced at  regular
 intervals.  Use of catalytic  purifiers  is  also
 effective (4).
  7.4 Column:
  7.4.1 Column  Tubing— For  most organic
 analyses in  aqueous systems, stainless steel is
 the  most desirable  column  tubing  material.
 However, when  analyzing organics that  are
 reactive with stainless  steel, glass  tubing
 should be used. With a flame ionization detec-
 tor, maximum resolution with packed columns
 is achieved  with  long, small-diameter (3.175-
 mm (0.125-in.) and smaller) tubing.  New tub-
 ing should be washed as described  in 6.1.2.
  7.4.2 Solid  Support—Maximum  column
 efficiency is obtained  with  an inert,  smail.
 uniform-size support. The lower limit of parti-
 cle size will be determined by  the allowable
 pressure drop across a column of given  diame-
 ter and length. Elimination  of  fines will  re-
 duce the pressure drop and  allow the use of
 smaller panicles: the commonly used size is
 80/100 mesh. Supports, which are not inert,
 may  cause  varying  degrees  of peak  tailing.
 Few supports can be classified as totally inert:
 however, techniques are available to assist in
 the deactivation of the support. Chromosorb
 "W",* the least active type of diatomaceous-
 earth support,  can be further deactivated by
 acid or base washing. A combination of acid
 washing and silanization (for example, dimcth-
 yldichlorosilanc (DMCS), hexamethyldisilane)
  '"Reagent  Chemicals,  American Chemical  Society
Specifications." Am. Chemical  Soc.. Washington.  D C.
For suggestions on the letting of reagents not listed by the
American Chemical  Society, see "Rcaeent Chemical! and
Standard!." h> Joseph Rosin. D. Van Nostrand Co., Inc..
New York. N.Y.. and the "United States Pharmacopeia."
  * This material, while proprietary in nature, a distinctly
superior to others which have been tried and is available
from essentially all vendors of chromalogrxphic supplies.
                                         483

-------
                                             O 2908
treatment may  reduce the surface activity still
further.  However, silanization  can decrease
column life. DMCS  treatment is particularly
useful when low bquid leads are used. Treat-
ment with specific chemicals that approximate
the properties of the sample being analyzed
has also  proven successful.  For example, ter-
cphthalic  acid treatment of Carbowax 20M*
reduces organic acid  and phenolic tailing. Use
of fluorocarbon supports can significantly re-
duce tailing. For low  boiling materials, porous
polymer  beads  formed by the polymerization
of monomers such as styrene wiih divinyi ben-
zene as a crosslinker  are finding more applica-
tion in trace analysis. Since there is no liquid
phase, there is  minimal column  bleed during
temperature programming.  In addition, elimi-
nation of the conventional solid support  re-
moves the adsorptive sites which  normallx
cause tailing. Caution must also be taken not
to exceed the recommended  maximum  tem-
perature  limit of the fluorcarbon supports or
of the porous polymer beads being used.
   7.4.3 Liquid Phases—Maximum resolution
and  minimum  baseline  noise  and drift are
achieved  with a relatively  lightly loaded col-
umn, (less than 5 percent) containing a stable
substrate of low volatility.  However, analysis
of aqueous samples with light column loading
produces shorter column life and a  greater
tendency for a shift in  retention  times and
delayed  elution as the column ages. Acceler-
ated aging will  occur if the  maximum temper-
ature limit of the liquid phase is exceeded or
approached repeatedly. Substrates should be
selected to permit operation at a temperature
below the maximum allowable if at all possi-
ble. Selection of liquid phases should be based
on the properties of the sample to be  ana-
lyzed. In general, polar substrates will resolve
polar compounds b> order of relative volatility
and polarity.  Polar substrates will  resolve
nonpolar comoounds by structural t>pe.  Non-
polar substrate:- »iil separate nonpolar  com-
pounds b\ volatility  and polar compounds by
structural  type, r-or examples  of applicable
liquid phases for a particular application, con-
sult published  methods for  specific  organic
classes.
   7.4.4  Column  Conditioning   All new col-
umns should be pre-conditioned to drive off
the  rcMdual contaminants which would foul
the detector and cause  severe baseline noise.
New columns can be conditioned by attaching
one end to the inlet port of the oven and al-
lowing 20 to 30 ml/min of carrier gas to pass
through the column either at 30 C above the
expected maximum operating temperature or
at the maximum temperature limit of the liq-
uid phase, whichever is lower.  The effluent
end of the column should be vented. The col-
umn should not be attached to the detector
during conditioning since  cluting organics
may foul  the detector. Occasional 5-drogen and air of
the highest initial purity which have been fur-
ther purified as described in 7.3. are fed to the
detector. Hydrogen can also be used which is
produced  from the electrolytic decomposition
of water.
  7.6 Glassware—All  glassware  that will
come into direct  contact  with the sample
should  be heated in  an  oven to  300 C
(overnight if possible) as a final cleanup step.
This will  serve to remove any source of or-
ganic contamination from prior work.
8. Samples and Sampling Procedure
  3.1  Sample Collection—Collect all samples
in accordance with the applicable method of
the American Society for Testing and Materi-
als as follows:
   D 510—Sampling Water,*
   D 1192—Equipment  for  Sampling Water
    and Steam.'and
  D 1496—Sampling  Homogeneous  Waste
    Water.*
Additionally  sample containers and  sample
MZC and storage shall be as specified in 8.2 to
8.4.
  8.2 Sample  Containers—CATC  should be
taken to collect  a  representative sample in a
clean, completely full glass bottle. The screw
cap should be  lined with aluminum foil or
TFE-fluorocarbon  to reduce the sorption of
                                           484

-------
                                             D2908
 insoluble organics.
   8.3 Sample Sire—The sample size must be
 small lo prevent overloading of the 3.175-mm
 (0.125-in.) columns generally used. For most
 aqueous analyses, a sample size of 2 to 5 u\ is
 generally  optimum. If the components of in-
 terest are of relatively high concentration, a I-
 >t\ sample is to be used. At low  concentra-
 tions,  a sample  approaching  10  «il  can  be
 used to increase the detectable limit although
 the measurement accuracy is slightly decreased
 since a 10-pl syringe is used. For the best ac-
 curacy, select a syringe with a capacity  50
 percent greater then the size of the sample to
 be injected.
  8.4 Sample Storage—Storage  time of sam-
 ples should be kept to a minimum.  If storage
 cannot be avoided, the bacterial action should
 be minimized by refrigeration, by  pH adjust-
 ment to about 2.0 (if organics  are not acid
 degradable). or by the addition of 1 ml of sat-
 urated mercuric chloride (HgCh)  solution to
 each liter of sample. Selection of a preserva-
 tion procedure is dependent on the analysis
 being made.

 9*. Preparation of Chromatograph
  9.1 Column—Select the  appropriate col-
 umn and install in the chromatographic oven.
 If the  column is new. it  should  be precondi-
 tioned  according  to  the  directions in 7.4.4.
 The column should then be attached to the
 detector and the  system checked  for leaks
 according to  6.1.3. The column  temperature
 requirements should be set  according to the
 requirements outlined in the specific method
 being used.
  9.2 Gases—With a flame  ionization detec-
 tor the gases require adjustment in the  ratio
 of about 1 part carrier gas to 1 part hydrogen
 to 10 parts air. A typical flow for the carrier
 gas when  using 3.175-mm(0.125-in.) tubing is
 25  ml/tnin.  Refer to  the  specific method
being used for flow requirements.
  9.3 Electrometer and Recorder—Adjust the
 electrometer and recorder as specified on the
instrument instructions so that the pen is ze-
 roed and the attenuation  steps  are  linear.
 Based on the organic content of the sample to
be analyzed, adjust the electrometer attenua-
tion  to give as near mid-scale deflections of
the recorder pen as is practical.
  9.4 Baseline Stability- Before proceeding
 with  the  analysis,  check  the stability of the
 recorder baseline with the pen at zero and the
 attenuation at the level  to  he used for the
 analysis.  If sporadic peaks occur, further col-
 umn  conditioning  may be necessary. The re-
 corder, electrometer, flow  controllers,  and
 flame detectors should also  be  checked as a
 possible source of the sporadic peaks.
   9.5 Column Storage— When  columns  are
 not in use. their ends should be capped. The
 need for reconditioning prior to  their reuse at
 a later time will be indicated by making cali-
 bration runs  with  a  known  concentration of
 standards. Reconditioning is generally  mini-
 mal  if the column  was  adequately purged
 prior to storage.

 10. Calibration and Standardization
   10.1  Qualitative:
   10.1.1 The  basic method  of tentative com-
 pound identification is by matching the reten-
 tion times of known standards suspected to be
 present with retention times of unknown com-
 pounds under identical operating conditions.
 The absolute  retention time is measured in
 minutes from  the time of injection to the peak
 maximum.  Since  retention  time  may  van
 significantly with concentration of the particu-
 lar organic compounds, identification is done
 more positively by spiking  the sample  with
 the suspected constituent and noting an in-
 crease in peak height.  In some instances more
 than  one  compound may elute  at the  same
 time  and   therefore have identical retention
 times. This condition  can often be recognized
 by a poorly shaped  peak (that is. double apex
 or  shoulder).  When  this occurs,  additional
 column(s) with different physical and chemi-
 cal properties  will be required to separate the
 combined peaks. An alternative, which is fre-
 quently preferable, is  to trap the peaks and
 identify them spectromctrically (see 11.7V
   10.1.2  Relative retention  times are devel-
 oped  by the insertion of a common noninter-
 fering organic into  each standard as well as
 into the unknown. The absolute retention time
of the common organic is then  divided  into
 the absolute retention time  of  each organic
 being analyzed. Utilization of relative reten-
 tion  times improves qualitative accuracy by
balancing  out numerous chromatographic var-
iations from run  to run.  for example, slight
variations   in column  temperature, program-
                                         485

-------
                                              D 2908
ming rate, carrier gas Row, and sample size as
well as column aging.
  10.1.3  Based on the type and concentration
of compounds expected in the sample to be
tested, prepare  similar standards in  reagent
water.
  10.1.4 At  least  three  relative  retention
times with a single column should be deter-
mined for each organic standard and the aver-
age  used  for qualitative analysis of  the  un-
known  sample. Relative  retention times
should be verified periodically.
  10.1.5  One  and two-column  identifications
are not usually sufficient for positive  identifi-
cation. A third column or spectrometric analy-
sis of the trapped peak will be required for an
unequivocal identification.
   10.2 Quantitative:
   10.2.1  The  quantitative  measurement  of
each component is  determined from the area
under the  individual chromatographic peaks.
Peak areas  can be determined  more  effi-
ciently by mechanical or  electronic  integra-
tors. If the peaks are symmetrical and sharp
with minimum  tailing, peak height   can be
used for estimating quantitative  response for
expediency in routine monitoring type analy-
sis.  The height is  measured  from the peak
maximum  to the baseline. If the peak occurs
in an area of baseline drift, approximate the
actual base of  the  peak  for measuring pur-
poses. Measuring  the peak width at one half
the peak height  and multiplying it by the peak
height will approximate the peak  area. The
error increases  as  the peak  width becomes
smaller or as peak tailing increases.
   10.2.2 Insertion  of an internal standard is
useful for quantitative analysis.  When  re-
sponse  is  calculated  relative to an  internal
standard,  compensation is provided   for  the
inadvertent changes  in chromatographic con-
ditions.  Selection  of the internal standard
should be  based on its separation from other
peaks, stability, and if possible on mid-chro-
matogram elution and structural  similarity to
the  components being analyzed. The  internal
standard should be applied at approximately
the  expected average concentration of the
organic constituents. When temperature pro-
gramming is used, two internal standards inny
be needed, one for  low-boiling  and  one for
high-boiling constituents.
   10.2.3 Mass response ratios are determined
by the  injection of standards containing  the
same concentration of both the internal stand-
ard and the individual components suspected
to be in the samples to be tested. For accurate
quantitative work  triplicate injections should
be made  on a conditioned column with  the
average being used  for further calculations.
All chemicals used should be of the highest
known purity, so that the appropriate correc-
tion may  be made when calculating the final
response factors. Response factors  should be
rechecked periodically.
   10.2.4 The linearity of the response factors
should be verified by varying the concentra-
tion of the individual  components over  the
concentration  range of interest while holding
the internal standard concentration constant.
These ratios when plotted  against concentra-
tion should yield  a straight  line that passes
through zero. Chromatographic operating
conditions should  always be  recorded on  the
graph. Attenuation should  preferably be  ad-
justed to keep the peaks at approximately 50
percent of full  scale,  if possible.  The final
peak areas or heights are  adjusted  according
to the electrometer attentuation setting used
for calibration.

11. Sample Testing Procedure
   11.1 Injection Practice—Use  a firm, rela-
tively fast injection  technique so that  the
sample can be injected either into the middle
of the injection port for flash vaporization, or
approximately 2 in. down  the column for  on-
column injection in a slug-like condition. Slow
injections may cause  poor resolution and
spreading. Use the same rhythm each  time.
Wash the syringe several times between  injec-
tions with acetone, then rinse with water, and
air dry by attaching to  a vacuum line.  Flush
the syringe at  least two times with the sample
to be analyzed. Remove the bubbles by pump-
ing the syringe plunger followed by a slow
drawup of the sample.  When injecting large
samples at high inlet pressure  (for example.
50 psi). hold the plunger so as to prevent its
blowout caused by the pressure buildup  in  the
injection port: special syringes are needed for
high-pressure work.
   11.1.1 Sample   Injection—Use   direct
aqueous injection  whenever  possible to pre-
                                           486

-------
                                               D 2908
 vent both the loss of some component-, and the
 introduction  of  extraneous peaks .that  may
 result  from concentration techniques  How-
 ever, when analyses are in the part per billion
 range,  concentration  techniques will  be re-
 quired. Carbon adsorption, gas stripping, sol-
 vent extraction, and freezeout  have been
 shown  to increase component concentration to
 detectable levels«IJS.6\.
   11.2  Establish operating conditions  identi-
 cal to those used for calibration and standard-
 ization. If changes are  required  because of
 sample peculiarities,  repeat calibration and
 standardization using the new conditions. If
 an internal standard is used, minor changes in
 operating conditions are tolerable.
   11.3  Inject sample  prior to  insertion of in-
 ternal standard to assist in either the selection
 of the internal standard,  or to assure that the
 internal standard selection  is well  resolved
 from component peaks in the sample. An open
 position in the chromatogram is selected for
 this purpose.
   11.4  Add  the  internal  standard(s) into the
 sample at  a concentration approximating the
 components  to be  analyzed  and  repeat  the
 analysis.
   11.5  Refer to  the specific method for sug-
 gested  sample size: 3 to 5 ul are often used.
   11.6  Determine  the  absolute retention
 times of the  individual  components in  the
 sample. Calculate  relative retention times
 using the retention time of the internal stand-
 ard  in  the denominator. Refer to the  pre-
 viously  developed listing for relative retention
 times of known compounds on specific col-
 umns:  if absolute retention times are  used.
 run  standards several times during  the test
 series.  Repeat on additional columns as neces-
 sary to  increase qualitative accuracy.
   11.7 Trap  individual  peaks for confirma-
 tory  analysis.  Mass spectromctric analysis of
 trapped components  is  often most informa-
 tive: however, infrared spectroeraphic  analy-
 sis, thin-layer chromatography. and microcou-
 lometry or  other speciali/ed  detectors  (for
 example,  flame  photometric detector,  modi-
 fied flame halogen detector) are iil Gas-
   Liquid Chromatoeraphv." International Journal
   of Air and  Water Pollution. UPWA, Vol 26.
   1966. pp. 591 to 602.
|J) Baker. R. A.. "Volatile Faiu Acids in Aqueous
   Solution by Gas-Liquid  Chromatoeraphx."
   Journal of Gas Chnjniaii>tmph\. JGCRA. Vol
   4. 1966. pp. 41$ to4|<)
|4> Baker. R. A. and Malo. B. A.. "Phenolic* b%
   Aqueous-Injection  Gas  Chromatographv."
   Journal of Environmental Science and Techftol-
   of\\ Vol I. 1967. pp. 997 to 1007.
(5» Baker. R. A.. "Trace Organic Contaminant
   Concentration by  Freezing-l;  Low Inorganic
   Aqueous  Solutions." Journal of the lmt*na-
   tional Association on Water Pollution Research
   Vol I. 1967. pp. 61 to 77.
t6t Baker. R. A.. "Trace Organic Contaminant
   Concentration  by  Freezing-11:   Inorganic
   Aqueous  Solutions." Journal of the Interna-
   tional  Association  on Water  Pollution  Re-
   search, Voll. 1967.pp.97to MJ.
                                          487

-------
                          NO. 4
NTTRILES IN AQUEOUS SOLUTION BY GAS-LIQUID CHROMATOGRAPHT

-------
          Designation: O 3371 - 74 T
         Tentative Method of Test for
         NITRILES  IN  AQUEOUS  SOLUTION  BY GAS-LIQUID
         CHROMATOGRAPHY1

 Thi* Tentative Method has been approved by the sponsoring committee and accepted by the Society in accordance with
 established procedures, for use pending adoption as standard. Suggestions Tor revisions should be addressed to the Society, at
 1916 Race St.. Philadelphia. Pa. 19103.
 1.  Scope
   1.1  This method covers nitriles that can be
 separated and detected quantitatively at a limit
 of approximately 1  mg/litre by aqueous injec-
 tion on a selected gas-liquid chromatographic
 column.
   1.2  This  method utilizes the procedures and
 precautions  as  described  in Recommended
 Practice D 2908.
 2. Applicable Documents
  2.1  ASTM Standards:
  D 2908 Recommended Practice for Measur-
     ing Volatile Organic Matter in Water by
    Aqueous Injection Gas Chromatography*

 3. Significance
  3.1  Nitriles at concentrations of a few milli-
 grams per  litre are potentially toxic to  aquatic
 life. Nitriles in waste water discharges should
 be detected and controlled.
  3.2  Gas-liquid chromaiography (GLC) can
 detect and  determine  mixtures  of nitriies  at
 levels  where wet chemical  procedures  are not
 applicable.

 4. Special  Comments
  4.1  It is recommended  that  samples that
cannot be  analyzed  immediately, be quick
 frozen  for  preservation. Samples should  be
 aeutralized to pH 7 at the time of collection to
minimize hydrolysis of the nitrite groups.
  4.2  Samples of nitriles to be employed as
standards should be considered to be unstable.
Storage in a freezer is recommended.
  4.3  It is  not  always  practical  to translate
operating conditions directly from  one GLC
instrument  to another. An operator  should
 optimize his instrument to a particular proce-
 dure, for example, injection and detection tem-
 perature, flow rates, etc.

 5. Typical Chromatograms
   5.1 The  following  instrument  parameters
 were used to obtain the typical chromatograms
 (See Figs. 1 and 2).
   5.1.1  Column—vs   in.  outside  diameter
 stainless steel, 8 ft long packed with a porous
 styrene divinylbenzene polymer.
   NOTE—"Chromosorb" 101. 50/60 mesh, was used
 for tbe typical chromatograms.
   5.1.2 Detector, flame ionization.
   5.1.3 Temperatures:
     Injection port              240*C
     Detector                  240*C
     Oven, isothermal            130*C
     Oven, programmed at        llO'Ciomax
       10'C/min                 of200*C
   5.1.4 Carrier Gas, helium at 25  ml/min.
   5.1.5 Sample Size:
               isothermal 5 *il
               programmed 3 u\
   5.1.6 Recorder, 3/i in./min chart speed and 1
 mV  full-scale response.
   5.2  Kovats Index Values:*
     Compounds
 Acetonitrile
 Acrylonitnle
 Relative
Retention
  1.00
Kovats
Index
 470
 512
  1 This method ii under the jurisdiction of ASTM Commit-
tee 0-19 on Water and is the direct responsibility of Subcom-
mittee D 19.05 on Inorganic Constituents in Water.
  Current edition approved Nov. 4, 1974. Published Feb-
ruary 1975.
  'Ann-jal Soot of ASTM Standards. Part 31.
  ' Gas  Chromatographic Data  Compilation,  ASTM
AMD 23A. Am. Sot Testing Mats.. 1967.
                                            534

-------
                                                     D3371
Propnoortnie
Iwjvtkronitnk
Vileronnnle
Hcuaenimlc
Beiuonitnle
                          1.67
                          2.1\
                          I.JO
                          304
                          3.JS
                          4.2i
                          5.42
570
635*
678
740'
783
90S*
990
6. Precision
   6.1  The precision of this method within  the
range from  10 to 60  mg/litre of standards in
distilled water may be expressed  as  follows:
      Compound
Acetonilnle
Propiominle
Meihoxy Acetonitnlc
Butyronitrile

where:
Sr
mg/ litre
           5r
Sr . O.OI5(mg/lure) -t- 0.9
5r - 0.088 (mg/litre) - 0.6
Sr - 0.097(mg/hlre) -r O.I
ST - O.IO(mg/!ilre) -0.4
                                                                  overall  precision, and
                                                                  concentration of the specific com-
                                                                  pound
                                                        ' Kov*u index vilues estimated  from relative retentioa
                                                     data became standard compound was not readily available.
                                               Tine IH
                                              J/*-(rleh a*r «inut«)
                          Column Packing - Oiromosorb IOI, 50/60 mesi
                          Corner Ga»- Helium at IS ml/mm
                          Temperature - Isothermal operation of the column at I30*C
                          Sample Size - 5 microlilen containing 10 mg/l of each niinie
                   FIG.
                                                  535

-------
                                                        03371
                                                                                                        3

                                                                                                        a
                                        nrraricii TI-C
                                    (Clwrt
                                                 l-iief> a..- iirv.t.1
                Column Packing • Chrotnosorta 101. 50/60 mob
                Carrier Gas- Helium at Z3 mi/min
                Temperature- Programmed operation at 10'C/min from 110'C to a maximum of 200*C
                Sample Size-} microliten containing l.iOO mg/lof each nitnle

                FIG. 2— PropoMd Tcaptncart CorooKognpbk ABaJyju of Ninila in Aqumo* Soivdo*


   Tkr Amtritm Societv far Testing aid Mattriais lakes no position respecting the validity of any patent ngnu asserted
in eonneeuon vuh any item mentioned in this standard. L'stn of this standard are expreaiy advised thai determination oftht
validity of any such patent nghu. and the risk of infringement of such rights, is entirety their own responsibility.
                                                    536

-------
                           CE NO.  5
DIRECT ANALYSIS OF WATER SAMPLES FOR ORGANIC POLLUTANTS WITH
            GAS-CHRCMATOGRAPHY--MASS SPECTROMETRY

-------
Direct  Analysis of  Water  Samples  for  Organic Pollutants with
Gas Chromatography-Mass  Spectrometry

Lawrence E. Harris, William L. Budde,1 and James W. Elchelberger
Envronmeniai Protection Agency. National Environmental Research Center. Methods Development and Quality Assurance ftose+rcft
laboratory. Cftcmnatt, Otuo 45268
A direct aqueous injection gas chromaiography-mass spec-
Irometry (GC/MS) procedure was explored as a supple-
ment to conventional solvent extraction for  analysis of the
organic pollutants in water and waslewater  samples. Stud-
ies were made of the effects of relatively large pressures of
water vapor on the well established electron impact frag-
mentation  patterns,  quadrupole  GC/MS system perfor-
mance, interactive background  subtraction, and detection
limits. It was concluded that  direct aqueous analysis is  a
valuable supplemental procedure for the detection of vola-
tile compounds that are not found with solvent extraction.
  Effective  water  pollution  control  requires  analytical
methodology that is capable of generating correct identifi-
cations and  measurements of the concentration of the or-
ganic pollutant? in water samples. This methodology is
necessary to determine the exact sources of pollution, to set
effluent standards for toxic pollutants, to enforce effluent
guidelines, to evaluate the effectiveness of treatment facili-
ties, and to  determine the causes of taste,  odor, and fish
kills.
  In the past, a very significant amount of research, fre-
quently over several weeks or months, was required to ob-
tain identifications of the trace organics in water samples.
Often this effort resulted in just a few identifications and it
occasionally produced erroneous  results. The development
of computerized gas chromatography-mass spectrometry
(GC/MS) revolutionized the field of trace organic analysis
(7,  2). Today many laboratories have the capability to
make more than a dozen unambiguous identifications with
just a few man-hours of effort.
  The first sample work->up methods used with GC/MS in
organic water pollutant  analysis were minor modifications
of standard  solvent extraction procedures which were de-
veloped for  pesticide analyses. These procedures together
with GC/MS are very effective in isolating, concentrating.
and identifying extraciable and volatile organic pollutants
at levels as low as 10 parts per trillion ilO ng/1). This great
sensitivity is achieved, in part, by an efficient concentra-
tion of a relatively large volume of organic solvent extract
to a very small volume. Concentrations of trace organics by
a factor of 10" is not uncommon.
  Solvent extraction? doe>. however, possess several limita-
tions including  the lo^s  of very volatile organic pollutant;
(e.C.. chlorinated solventsi by vaporization during the ex-
tract concentration step. Another difficulty is the failure t<>
extract efficiently a varietv of volatile but water soluble or-
ganic pollutants ie s . low molecular weight alcohols and
ketone solvents).  A supplemental work-up  procedure  for
   1 To whom correspondence should be addressed.

(1) ft. A. HrtesandK 8«mann. Sdf.tca. 178, 158(1972).
(2) J. A" EicneOcfcer. L. £. warns, and VV L Budde. Antl. Chem.. 46. ?rT
   (1574).
the analysis of these compounds is required. Perhaps the
simplest and most direct approach is the analysis of the un-
altered water samples by GC/MS.
  The gas chromatography of unaltered water samples is
feasible and has been known for some time (3-7). It  has
been  practiced only sparingly, however,  because  conven-
tional GC detectors (e.g.. the flame ionization detector) do
not produce sufficient information to  unequivocally distin-
guish among  the  enormous variety  of different  organic
compounds that could be present in a water sample. Com-
puterized GC/MS overcomes this difficulty since the mass
spectrometric  data are frequently sufficient for an unam-
biguous characterization of most of the very volatile com-
pounds present.
  Routine direct  aqueous GC/MS  analysis  for  organic-
water pollutants offers the potentially very significant ad-
ditional benefit of instant  analysis. Since no time and
labor consuming pre-analysis treatment is required, a rela-
tively large number of samples may be processed per unit
of time at a relatively low unit cost.
  A study was made of the applicability of this technique
to water pollutant identification.
  Difficulties that might be anticipated  with water as a sol-
vent for GCA1S analysis were studied  also. Since  the sol-
vent extract concentration step was eliminated, sensitivity
limitations were defined. Traditionally water is considered
highly detrimental in magnetic deflection mass spectrome-
ters. It may facilitate discharges from 2-8 kV accelerating
potentials and cause degradation of Cu-Be electron multi-
plier detectors. However the effects of large quantities (1-
10 pi) of  water injected  into the GC/MS on the  perfor-
mance of the quadrupule mass spectrometer and the sam-
ple enrichment device were  unknown. Also unknown was
the effect of large quantities of water  vapor on the  well es-
tablished electron impact fragmentation patterns of organ-
ic compounds.
  The approach used was to analyze  representative waste
samples and well-defined mixtures of compounds to ascer-
tain the effect of water on the system  and fragmentation
patterns. The  levels of detection of a variety of classes of
compounds were determined with several new and  old GC
column packing materials.

                  EXPERIMENTAL
  Instrumentation. The water samples were analyzed using di-
rect on-column injection into a Varian Model 1400 gas chrnm.ito-
graph  interfaced with a Kinnipan Model 1015 C quadrupole moss
spectrometer system controlled by a Systems Industries 120 data
acquisition and control system (2). The MS wat the only  detector
used. The data were displayed as plots on  a cathode ray tube dis-
(3) R. C Dress/nan. J. Chromarogr. Sci.. 8. 265 (1970).
(4) F  I. Onuska. Water Res.. 7. 835(1973).
(5) M. E. Foi. Snvron. Sci. Technol.; 7, 838 (1973).
(«) J. W. Sugar and R. A. Conway. J. w*ier PotM. Contr. Feet, 40, 1622
   (1968).
(7) "Annual Book of ASTM Standards." ASTM. Part 23. Philadelphia, Pa., c
   1972. pp 706. fl 19.
1912 • ANALYTICAL CHEMISTRY, VOL. 46. NO. 13, NOVEMBER  1974

-------
Table I. Selected Organic


Compound
Blank
n-Decane
n-Decane
MIBK
MIBK
n -Butyl acetate
n -Butyl acetate
M-Amyl alcohol
M-Amyl alcohol
p-Cresol
^-Cresol
Acetophenone
Acetophenone
2-Phenylethanol
2-Phenylethanol
w-Hexadecane
w-Hexadecane
sec -Butyl alcohol
Acetone
Methyl -n -octanoate
Chloroform
DME
H-Amyl alcohol
M-Amyl alcohol
'Methylene chloride
Methylene chloride
Ethyl acetate
DME
MIBK
Dioxane
Acetophenone
o-Chlorophenol
w-Cresol
Compounds
Puutltv

•9

100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
20
5
10
10
5
50
50
50
50
50
50
50
Analyzed by CC/MS
Quantity
•olvcot.
»L
1
1
1 •
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
2
1
10
10
1
1
1
1
1
1
1
1



Sohrat
Water
Acetone
Water
Acetone
Water
Acetone
Water
Acetone
Water
Acetone
Water
Acetone
Water
Acetone
Water
Acetone
Water
Water
Water
Water
Water
Water
Water
Water
Water
Water
Water
Water
Water
Water
Water
Water
Water
                                                                                   Concentration,
                                                                                      019/L
                                                                                       100
                                                                                       100
                                                                                       100
                                                                                       100
                                                                                       100
                                                                                       100
                                                                                       100
                                                                                       100
                                                                                       100
                                                                                       100
                                                                                       100
                                                                                       100
                                                                                       100
                                                                                       100
                                                                                       100
                                                                                       100
                                                                                       100
                                                                                       100
                                                                                       100
                                                                                       100
                                                                                        10
                                                                                         5
                                                                                         1
                                                                                         1
                                                                                         5
                                                                                        50
                                                                                        50
                                                                                        50
                                                                                        50
                                                                                        50
                                                                                        50
                                                                                        50
                                         cc
                                       column

                                          1
                                          1
                                          1
                                          1
                                          1
                                          1
                                          1
                                          1
                                          1
                                          1
                                          1
                                          1
                                          1
                                          1
                                          1
                                          1
                                          1
                                          1
                                          1
                                          1
                                          1
                                          1
                                          1
                                          1
                                          1
                                          1
                                         2
                                         2
                                         2
                                         2
                                         2
                                         2
                                         2
CC column
condition)
    a
    b
    b
    b
    b
    b
    b
    b
    b
    c
    c
    c
    c
    c
    c
    c
    c
    d
    d
    d
    a
    a
    a
    a
   a
   a
   e
   c
   e
   e
   e
   e
   e
   « 100" isothermal;' 70* for 2 min. then 6*/min to 120*;' 150* for 2 min; then 6*/min to ISO": •* 100* for 1 min. then 6*/cnin to 150'; • 60* for
 3.5 min, then 12' /min to
 play unit (Tektronix Model -1010) or a flat-bed plotter (Houston
 Instruments  Model  DP-1). The GC/MS interface utilized an all
 glass jet-type enrichment device to deliver the sample directly into
 the ion source of the MS. The butch iniet system was all glass with
 a constant-leak  opening that introduced the sample directly into
 the ion source of the MS. The  batch inlet was  heated to 100° for
 the analyses. Other conditions that were held constant throughout
 the analyses  were: helium carrier gas  at a flow rate of about 30
 ml/min; temperature of the GC injection port at 190"; the temper-
 atures of the  interface and transfer line at 210°; detector manifold
 temperature at 100": pressure in MS of 10-i Torr; ionizing voltaee
 of 70 eV; A filament current of 500 uA: electron multiplier at .1000
 volts; mass rar.pe scanned  from 'JO-'JOO amu at an integration time
 of 10 msec.'smu: and sensitivity at 10~7 A/Volt unless otherwise
 specified. The quadrupolc MS operating parameters were adjusted
 to give the normal ion abundances for a standard  reference Com-
 pound (S).
  Gas Chromntocraphv  Columns. Column ]. .\  12-foot coiled
 stainless steel lo.d. ll.ll.'"> in.) tulxr WHS parked with (>0'SO mesh CAS
Chmwn Q ctiulrd with .v\. CartHiwax ?lk M.
  Column 2.  A 7-loot coiled class column  li.d.  2 mm) was packed
 with 60/80 mesh arid-washed Chromusorb \V con ted with 10% Free
 Fatly Acid Plui,.-(FKAIM.
  Column •>'. An S-linil foilrtl ^l:i>» o'lumn (i.<). 'J mm) was parked
with .Ml/liO nif>h t'hrmuuMirli 101. Sfirridl  care was lakrn to l',u>h
the column sullicienily with helium i-lO-"><) min) to rrmnve any re-
sidual air faun the |xu-kmi; Iwlore heat inc.
  Method I.  An ur£:mir i»iu|Hiuiid tl  «1) WHS  va|mrizvd into the
          rt-srrvtMr MIH] ulU'^iii lu lc;ik ultiwly ihriiuxh a ^laxs Irit
directly into the ion source of the mass spectrometer. The mass
spectrum of the compound was acquired repetitively trum 20 to
200 amu using computer control, and the data were stored on a
disk storage device. As  the  MS data were being generated and
stored. 1 »d of tap water was injected onto GC Column 1 at 100°.
After the water eluted from the column into the detector and was
pumped  out of the MS. valves were opened and the organic com-
pound was pumped from the batch inlet. The experiment was ter-
minated, and the data were recalled for evaluation  of the mass
spectra.  The organic  compounds that  were analyzed using this
method include the following:  1.2-dimethoxvethane (DMK). di-n-
butyl amine. svc-butyl alcohol, methyl-n-octanoate. acetic acid, n-
hexadecane. n-amyl alcohol, and n-butyl acetate.
   Method 2. U'ater 11 n\) was vaporized into the batch inlet reser-
voir and allowed to leak continuously into the detector. Then 1 ul
of acetone containing 100 ng each of n-decane, methyl-IM.butyl ke-
tone (MIBK). n-butyl acetate, and /i-amyl alcohol was injected
onto Column 1 at 70°. After the ncetone solvent was  pumped out
of the detector, the ionizer was turned on and the mn.ss spectra of
the organic compounds were repetitively scanned whilo water was
continuously leaking into the ion source. The GC temperature was
programmed from 70 to IVJO" jit 6°/min.
  Method 3. The m.»s spt-ctra ul selected organic compounds
were measured liy injcv'tm" aquivnis :iiul/ur.ai-rtiiiu* >«>liili\in» into
the (iC/MS. The roni|hiumls sturiifd. i|iiantilu-s. Milvrnls. .u»l (iC
conditions arc shown in Tahlc 1.
  Waste Effluent Sample. This s;impl«- w;ts n<-()uin-(l from the ef-
llurnt  of a la>:iN>ii typ«- flu-mical trcaliiu-nl  l.u-ility and '.vj> ana-
lyzed  by direct.injectiun ul 1 n\ IMIIO i'uluinn '.i. The nilumn tcm-
                                                          ANALYTICAl. CHEMISTRY.  VOL. 46. NO. 13. NOVEMBER 1974 • 1913

-------
                                                                  Butyl Act late
                                                                      From me
                                                                    Batch Inlet
                                                                                      DT—sj
                                                                                 Background
                                                                                 in  System
                                18	23   33   KJ   S3   60   78  83   33   100  118  123  130  110
                                SflECTH/l H/BEH
Figure 1. An ion aburviance ctvomatogram obtained from water eluting from Column 1 while n-butyl acetate was permitted to leak continuous-
ly into the ion source from the batch inlet
perature was held at 120* for 3 min, then programmed to 180° at
l2Vmin.
  Tap Water Sample. A 5-«l  sample of tap water was injected
onto Column 3 at 140°. Ions of mass 47, S3. and So were repetitive-
ly observed with an integration time of 450. 900, or 1350 msec
which was determined dynamically  to maximize signal to noise.
The technique of subset data acquisition was described previously
(2). The  technique of integration time  a» a function of signal
strength will be described in the near future ie>. Additional details
on the analysis of chloroform in tap water will be presented in a
future paper (9).

            RESULTS  AND DISCUSSION
  The mass spectra of a number of organic compounds
were recorded while water was present in relatively high
proportion in  the ion source. The purpose of these experi-
ments was to determine the  effects of water on well  known
fragmentation patterns. Method 1 was used to study the ef-
fect of an increase in water pressure  from  the GC  inlet
(Column  1) on the fragmentation patterns  of several com-
pounds which were permitted to leak continuously from
the batch inlet. For example, mass spectra of n-butyl ace-
tate were acquired and  recorded continuously during one
experiment and  Figure  1 shows the ion  abundance chro-
matogram that was obtained. In regions I, II, and  III, n-
butyl acetnte  spectra  were  acquired. In region II, water
eluted from the column and  subsequently was pumped out
of the mass spectrometer. .In region IV, the  batch inlet sys-
tem valve w.i$ closed and n-butyl acetate wns pumped from
the mass spectrometer. From this experiment, mass spectra
of n-butyl acetate were plotted before (spectrum 60), dur-
ing (spectrum  75), and  after (spectrum  So)  water  eluted
from  the  GC  inlet system. Spectrum 130 was subtracted
from  each spectrum to  correct for background  ions. The
masses and abundance data  from these spectra are  shown
in Table  II. For  any given  ion, the relative abundance is
netrly identical in all  three  spectra,  which indicated  that
xvnicr had no  significant effect on the fragmentation  pro-
ct-sses of this compound.

(8) J W. Eicfvritiergof. 1. E. Harris, and W. U Budda. presented at trie 22nd
   annual c.-nlci^ce on mass spoclronwuy 0nd >lto>d loptcx. Pnrtadetphia.
   Pa.. May 13-24. 1974. In press.
(9) T. A. Bdi.ir and J J. Ucriteruxxg.  Environ. Sd Tocfinot.. In pr«sa.
 Table II. Mass Spectra of Butyl Acetate from Batch
     Inlet with Water Eluting from GC
                          Relative ion abundance*
m/«
39
40
41
42
43
44
45
55
56
57
58
61
71
73
87
Spectrum
No. 00-130
5
1
17
5
100
. 3
2
7
35
8
2
12
2
12
2
Spectnsn
So. 75-130
5
0
19
4
100
4
2
7
32
8
2
12
2
12
2
Spectrum
JCoVSS-130
6
1
19
6
100
3
2
7
36
9
2
13
2
14
2
   Similar results were obtained with the other compounds
studied by Method 1. These compounds included methyl
n-octanoate which has a base peak  in its mass spectrum
that is due to a McLalTerty rearrangement. In this well
known  process, a gamma hydrogen  is transferred  to the
carbunyl oxygen and a neutral Cr.Hia fragment is expelled
from the molecular ion. It was clear that the relatively high
pressure of water  present during the  ionizalion did  not
cause ion-molecule  reactions or other  effects that  would
alter this fragmentation process. We concluded that the ion
source design, pumping speeds, etc. that were employed
were such that disruptive effects were precluded in general.
  This  conclusion  was confirmed by several experiments
which  used Method  2 to study the effect of a constant
water pressure from the batch inlet on the fragmentation
p.ittern of several compounds introduced from the (  H«-U/C»JOCO

-------
                      !„
                      i.
                      t.
                      f.
                      r
                     5*.
                     *..
                      K.
                      «.
                      t.
   u-4
   ISMKt
   1011
;  I
   I  •
                                                                   f» !• US «i B» IB to <••
 figure 2. An ion abundance crtromatogram Obtained from a mixture of SO ng of each of the following compounds in tap wafir on Column 2
 (1) carton dioxide tram the tap water sotwenc (2) etnyl acetate: (3) 1.2-dimetnoxyethane: (4) methyl isobutyt ketone: (S) Ooxane: (6) waier [7> •cetoptenone: (8)
 »cMorophenot: and (9) /rvcrtsol
 time the compounds were undergoing ionization and frag-
 mentation. One of the compounds studied was n-amyl alco-
 hol which undergoes an electron impact induced dehydra-
 tion reaction. In this and all other experiments the pres-
 -nce of water in the MS detector system caused no notice-
 able effect on the observed mass spectra.
   A  number of additional experiments (Method  3) were
 conducted to support these conclusions. The  compounds
 shown in Table I were introduced into the mass spectrome-
 ter through the GC inlet. The compounds were selected as
 representatives of several classes of compounds commonly
 found as voiaciles in water samples. The mass spectra that
 were obtained from acetone and water solutions were com-
 pared and found to be identical in ail cases.
   Precautions and System Performance. In principle,
 the  sample  enrichment  device in the  GC/MS interface
 should reduce the amount of water which  enters the ion
 source of the mass spectrometer. Although no quantitative
 measurements were made,  our qualitative observations
 support this expectation. As much as 10 u.1 of water was in-
 troduced  onto the GC  column in a single injection  and a
 number of 1- to 10-»il injections were made  during a work
 day with no apparent detrimental effects on  system perfor-
 mance or sensitivity.
   With cross-linked porous polymer packed columns, e.g.,
 column 3, from which  water elutes  very quickly,  the  ion
 source potentials (5-100 V) and electron multiplier voltage
 (3 kV) were not applied and data acquisition was not begun
 until water  eluted. This is also standard procedure tor or-
 ganic solvent extracts. With other columns, e.g.. Columns 1
 and 2. from which water eiutes much later and after some
 organics. the source and multiplier potentials were applied
 immediately after injection and left on during solvent eiu-
 tion with small,  i.e., 1 n\. water injections. With larger
 quantities, these potentials were usually removed during
 elution of the  water (Figure 2). A downward drift in the
 ionizing current was observed with the ion source on while
 1 ill or more  of water eiuted (Fiirure 1).
  During the course of these experiments over more than
 IS months, frequent observations were made of overall sys-
 tem  performance. The jx>rformance measurement  was the
 ability of the system to produce, from a '_'b-ng injection, the
 furrifrt electron impact fragmentation pattern  of  a refer-
 ence compound US)  with u molecular weight of -442. The
 background  inn's* was never observed at jrre-Jter than '2-;J"»»
of the base peak. Normal  degradation in the performance
nf a qiKiiirupole mass .spectrometer is revealed by a loss in
M-iisitivity (signal/noise) and resolution ut musses  greater
than •—2.~»0 amu. This is caused frequently by an accumula-
tion  of carbon and other  extraneous  deposits on  the ion
source and i|u:uirupole rods. Surprisingly, it was our quali-
                          tative observation that normal performance degradation
                          was retarded  somewhat during the  period  of intensive
                          study of aqueous injections. We tentatively credit this ap-
                          parent effect to a steam cleaning phenomena.
                            Background Subtraction. Care must be taken during
                          computer assisted  background subtraction  if the well es-
                          tablished fragmentation pattern? are  to be observed. For
                          example,  n-decane was  chromatographrd  in  acetone and
                          water on Column  1 (Table I). The base peak  of n-decane.
                          after background subtraction, in acetone wa* mass 57 and,
                          in water,  mass 43.  The background spectrum selected for
                          subtraction in each case was immediately !n front of the n-
                          decane peak. Acetone has a base peak of m;iss 43 and resid-
                          ual acetone in the spectrometer contributed a large mass 43
                          ion to the background spectrum. Therefore,  the mass 43
                          ion abundance was reduced  substantially  by background
                          subtraction in the  spectrum of n-decane in acetone, and
                          mass 57 became the base peak. This did not occur in water
                          and is a clear advantage of a  non-organic solvent for GC/
                          MS analyses of organic pollutants.
                            However, an ion of mass 44 was observed  as the  base
                          peak for a tap water blank on Column 1 (data acquisition
                          began  at mass 20). This was not caused  by dissolved carbon
                          dioxide, which generally elutes as a sharp peak (Figure 2).
                          It may emanate from the continuous decomposition of car-
                          bonates or other dissolved compounds, n-Butyl  acetate in
                          water  eiuted from  Column 1  on  the leading  edge  of the
                          water  peak. Subtraction of a background spectrum from
                          before the n-butyl acetate peak save a  spectrum  similar to
                          that of n.-butyl acetate but with mass 44 as the base peak.
                          If a spectrum  for subtraction was  chosen  from near the
                          apex of the water peak, the mass 44 was eliminated and the
                          correct n-butyl acetate spectrum was obtained. This illus-
                          trates  one  necessary precaution of aqueous injection  GC/
                          MS. The ability to rapidly (10-15 sec) view a mass spec-
                          trum histogram on a CRT display greatly facilitates back-
                          ground subtraction and other type* of real time interactive
                          data reduction.
                           Column Selection.  The selection of a GC  column for
                          aqueous GC/MS  analysis depends <>n  the  types of com-
                          pounds sought  in the analysis. If it is desired to search for
                          low molecular weight volatile compounds, porous polymers
                          (e.g.. Column ;!) appear to be the best choice lor a column
                          packing material. If the aqueous analysis is to be extended
                         into the types of compounds usually sought !n- solvent ex-
                         traction, f.fi.. phenol and substituted phrnnls. .mother  col-
                         umn is required because elution. tjtws lor higher molecular
                         \v«-i;jht compounds become unacceptable. Kiiher I'arhowax
                         (Column  1) ur  KFA1*  (Column 2) si.itionarv |ih:i*fs are a
                         reasonable choice. The disaiiviini.i^es ol these  columns in-
                         clude the inability to observe, as JLstinot peaks, compounds
                  ANALYTICAL CHEMISTRY. VOL. 4&  wn
                                                                                                     1971 . .«..*

-------
                    1}
                    *'.
                    If.
Figure 3. An ion aOundance chromatogram obtained from a direct injection onto Column 3 of effluent from a lagoon chemical treatment facility

The compounds ocntihetf weft: (1) metfunot (2) etrunol: (3>acetone: (4) 2-vooanoi: (51 acetic acid; (6) 2-butanone: (7) propanoic acid: (8) 2-«moryetfiano<; (9)
          . i tQi inetriyi armnc: iiii M.AM>n«(nylfermanrude, and (12) pentanoic acd
       TR*
                                                ira  110
 Figure 4. ion aouncance cnromatogram obtained from tap water
 ekmng from Column 3 witn suaset data acquisition at masses 47,
 83. and 85
which have  the  same  retention time  as  water. Also, as
pointed  out  previously, it may be necessary to interrupt
data acquisition curinz eiuuon of water with these columns
(Figure 2). None of the numerous specialty phases avail-
able for  specific analyses were evaluated for this applica-
tion.
  Application to an Environmental Sample.  Numerous
water samples were collected  trorn  waste effluents, chemi-
cal  spill.-, and  waste treatment plants and  submitted to
aqueous  inject inn  CiC  MS. The inn abundance  chromato-
gram shown  in Figure  •'? wa.- uhtained I'rtim an aqueous in-
jection of the effluent from a lagoon type chemical waste
treatment plant. From each peak of the chromatogrnm. a
mas.s spectrum \va> retrieved Irom the disk storage device.
This provided an  unequivocal identification of  the pollu-
tants .-till present in the effluent.
  Detection Limits. The absolute detection limit of a qua-
drupole CC/MS system will vary widely and depend on a
variety uf factors including the efficiency of the GC column
and enrichment device, the presence of carbon deposits on
the ion source or rods, the adjustment, of the ion source and
rod potentials, the mass range scanned,  the integration
time per mass unit, the U>lal scan time, and the signal-lo-
noise ratio required in any given mass spectrum.  During
the course of these experiments with the quadrupole GC/'
MS, it was possible to obtain a reasonably clean (signal/
noise = 25 or greater) 40—{00 amu mass spectrum in a 5-sec
total scan time from about 5 ng of a volatile organic com-
pound. A 5-jil aqueous  solution containing a total of 5 ng of
a compound has a concentration of 1 mg/L (1 ppm) and this|
should be the approximate lower detection limit for a 40-
400 amu 5-sec scan. Using a somewhat shorter mass range.
(20-200 amu). methylene chloride and n-ainyl alcohol were
mixed in water at the 1 mg/L concentration and chromato-
graphed using Column  1 (Table I). Acceptable mass spectra
were obtained from this experiment. However,, with the po-
rous polymer packed  column (Column  3), the detection
level was about 10-20  rag/1. In general, the concentration
required to obtain a reasonably clean MS was between 1-50
mg/L This detection limit for conventional data acquisition
is usually not sufficient for relatively clean water, i.e., fin-
ished tap wacer or surface waters. However, it is more than
adequate for the analysis  of effluents and other relatively
dirty water which frequently contains organic compounds
at a concentration greater than 1 mg/1.
  There  are several methods available which may be used
to extend the detection limit. Very large samples, i.e., 50-
100 jil may be injected  if a solvent venting valve is installed
and a column is used from which water eiutes rapidly, e.g.,
Column  3. Another approach  to  enhance sensitivity  for
specific compounds utilized data acquisition from subsets
of the ions which are observed in conventional mass spec-
trometry (2). With this approach only a relatively few ions
are monitored in real time with a relatively long integration
time on each to substantially Improve signal/noise by time
averaging. Figure 4 shows a direct aqueous analysis of 5 n\
of a finished  tap water sample. Ions of mass 47, S3.  and 85
were  repetitively monitored with an integration time  of
450-1350 msec  which  was determined dynamically as a
function of signal strength. These ion* were  .selected be-
cause they represent the three most abundant ions in the
mass spectrum of chloroform. The peak observed had the
retention time  of chloroform  which was  estimated  to be
present in the 80-120 jtg/1. concentration range;.

                   CONCLUSION
  Direct aqueous injection - gas chromatogTaphy  with a
computer controlled quadrupole mass spectrometer detec-
tor  is a powerful supplemental method for the unambig-,
uoii£ identification of the  more volatile organic pollutants
in water sample*. Hela.tively large pressures of water vapor
in the mass .spectrometer have no significant effect  on the

-------
weii established 70-«V electron impact frn::montalion pat-    This technique made the mHlwwl applicable to the analysis
terns of organic compounds or the perfurniancc of the qua-    of relatively clean surface »r n'uikinp water.
drupoie GC/MS  system.  The detection limits attained
using conventional data acquisition wer«? l-SO ppm which                   Aownwi  •-nr \TPMT
makes the technique compatihle vith the concentrations of                   AC*U>U»> i-«.i;u.ne.r« i
organic compounds found in domestic sewage and  other      We thank Torn Bellar of th:.« laboratory for the original
waste effluent water samples. Greater sensitivity, to about    suggestion to analyze lap WHUV for chloroform.
50 ppb, was attained with real time data acquisition from
subsets of the ions used in conventional mass spectrometry.    RKCKlVF.lt for review May 8. 1974. Accepted July 19,1974.

-------
                         CE NO.  6
DETERMINATION OF ORGANOCHLORINE PESTICIDES IN INDUSTRIAL EFFLUENTS

-------
   1.      METHOD FOR ORGANOCHLORINE PESTICIDES IN INDUSTRIAL EFFLUENTS


  1.  Scope and Application


      1.1  This method covers the determination of various organochlorine


           pesticides, including some pesticidal degradation products and related


           compounds in industrial effluents.  Such compounds are composed of


           carbon, hydrogen, and chlorine, but may also contain oxygen, sulfur,


           phosphorus, nitrogen or other halogens.


      1.2  The following compounds may be determined individually by this method


           with a sensitivity of 1 ug/liter:  BHC, lindane, heptachlor, aldrin,
      M
      I  |
           heptachlor epoxide, dieldrin, endrin, Captan, DDE, ODD, DDT, methoxy-
        •
      -p
      P*   chlor, endosulfan, dichloran, mirex, pentachloronitrobenzene and tri-


   ~ ^   fluralin.  Under favorable circumstances, Strobane, toxaphene,


           chlordane (tech.) and others may also be determined.   The usefulness


           of the method for other specific pesticides must be demonstrated by


-—  *3      the analyst before any attempt is made to apply it to sample analysis.
        •
     ;!.«?  When organochlorine pesticides exist as complex mixtures, the


         .  individual compounds may be difficult to distinguish.  High, low,  or


        P=i  otherwise unreliable results may be obtained through misidentifica-


           tion and/or one compound obscuring another of lesser concentration.


           Provisions incorporated in this method are intended to minimize the


           occurrence of such interferences.


      Summary


      -.1  The method offers several  analytical alternatives, dependent on the


           analyst's assessment of the nature and extent of interferences and/or


           the complexity of the pesticide mixtures  found.   Specifically,  the


           procedure describes the use of an effective co-solvent for efficient


           sample extraction; provides,   through use of column chromatography
~  a
<
  _  "co"
^S,

-------
                                     1-2




         ana  liquid-liquid partition, methods  for elimination of non-pesticide




         interferences  and the pre-separation  of pesticide mixtures.   Identifi-




         cation  is made by selective gas chromatographic separations  and may




         be corroborated through the use of two or more unlike columns.




         Detection and  measurement is accomplished by electron capture,  micro-




         couloraetric or electrolytic conductivity gas chromatography.   Results




         are  reported in micrograms per liter.




    2.2  This method is recommended for use only by experienced pesticide




         analysts or under the close supervision of such qualified persons.




3.   Interferences




    3.1  Solvents, reagents, glassware, and other sample processing hardware




         may yield discrete artifacts and/or elevated baselines causing




         misinterpretation of gas chromatograms.  All of these materials must




         be demonstrated to be free from interferences under the conditions




         of the analysis.  Specific selection of reagents and purification of



         solvents by distillation in all-glass systems may be required.




         Refer to Part I, Sections 1.4 and 1.5, (1).




    3.2  The interferences in industrial effluents are high and varied and




         often pose great difficulty in obtaining accurate and precise




         measurement of organochlorine pesticides.  Sample clean-up procedures




         are generally required and may result in the loss of certain organo-




         chlocine pesticides.  Therefore, great care should be exercised in




         the selection and use of methods for eliminating or minimizing,




         interferences.   It is not possible to describe procedures for over-




         coming all of the interferences that may be encountered in industrial




         effluents.

-------
                                   1-3






    3.3  Pol/chlorinated Biphenyls (PCB's) - Special attention is called




         to industrial plasticizers and hydraulic fluids such as the PCB's




         which are a potential source of interference in pesticide analysis.



         The presence of PCB's  is indicated by a large number of partially




         resolved or unresolved peaks which may occur throughout the entire




         chromatogram.  Particularly severe PCB interference will require




         special separation procedures (2,3).




    3.4  Phthalate Esters - These compounds, widely used as plasticizers,




         respond to the electron capture detector and are a source of inter-




         ference in the determination of organochlorine pesticides using




         this detector.  Water leaches these materials from plastics, such




         as polyethylene bottles and tygon tubing.   The presence of phthalate




         esters is implicated in samples that  respond to electron capture but




         not to the microcoulometric or electrolytic conductivity halogen




         detectors or to the flame photometric detector.




    5.5  Organophosphorus Pesticides - A number of organophosphorus pesticides,




         such as those containing a nitro group,  eg, parathion,  also respond




         to the electron capture detector and  may interfere with the determina-




         tion of the organochlorine pesticides.   Such compounds  can be




         identified by their response to the flame photometric detector



4.   Apparatus and Materials




    4.1  Gas Chroraatograph - Equipped with glass  lined injection port.



    4.2.  Detector Options:




         4.2.1  Electron Capture - Radioactive (tritium or  nickel 63)



         4.2.2  Microcoulometric Titration




         4.2.5  Electrolytic Conductivity

-------
                                1- 4





 4.3  Recorder - Potentiometric strip chart  (10 in.) compatible with



      the detector.




 4.4  Gas Chromatographic Column Materials:




      4.4.1  Tubing - Pyrex (180 cm long x 4 ram ID)




      4.4.2  Glass Wool - Silanized




      4.4.3  Solid Support - Gas-Chrom Q (100-120 mesh)




      4.4.4  Liquid Phases - Expressed as weight percent coated on




             solid support.




             4-.4.4.1  OV-1,  3%




             4.4.4.2  OV-210, 5%




             4.4.4.3  OV-17, 1.5% plus QF-1, 1.95%




             4.4.4.4  QF-1,  6% plus SE-30, 4%




 4.5  Kuderna-Danish (K-D) Glassware (Kontes)



      4.5.1  Snyder Column - three ball (macro) and two ball (aicro)




      4.5.2  Evaporative Flasks - 500 ml




      4.5.5  Receiver Ampuls - 10 ml,graduated




      4.5.4  Ampul Stoppers




 4.6  Chromatographic Column - Chromaflex  (400 mm long x 19 mm ID) with



      coarse fritted plate on bottom and Teflon stopcock; 250 ml reservoir




      bulb at top of column with flared out funnel shape at top of bulb - a




      special order (Kontes K-420540-9011).




 4.7  Chromatographic Column - pyrex (approximately 400 mm long, x 20 mm ID)




      with coarse fritted plate on bottom.



 4.3  Micro Syringes - 10, 25, 50 and 100  ul



 4.9  Separatory Funnels - 125 ml, 1000 ml and 2000 ml with Teflon stopcock.




4.10  Blender - High speed,  glass or stainless steel cup.

-------
                                     1-5

    4.11  Graduated cylinders - 100 and 250 ml

    4.12  Florisil - PR Grade (60-100 mesh); purchase activated at 1250 F

          and store in the dark in glass containers with glass stoppers or

          foil-lined screw caps.   Before use, activate each batch overnight

          at 130 C in foil-covered glass container.  Determine lauric-acid

          value (See Appendix I).

5.   Reagents, Solvents, and Standards

    5.1   Ferrous Sulfate - (ACS)  30% solution in distilled water.
                          *
    5.2   Potassium Iodide - (ACS) 10% solution in distilled water.

    5.3   Sodium Chloride - (ACS)  Saturated solution in distilled water

          (pre-rinse NaCl with hexane).

    5.4   Sodium Hydroxide - (ACS) 10 N in distilled water.

    5.5   Sodium Sulfate - (ACS) Granular,  anhydrous (conditioned @ 400 C  for  4  hrs).

    5.6   Sulfuric Acid - (ACS) Mix equal volumes of cone.  H.SO.  with

          distilled water.

    5.7   Diethyl Ether - Nanograde,  redistilled in glass,  if necessary.

          5.7.1  Must contain 2% alcohol and be free of peroxides by

                 following test:   To  10 ml  of ether in glass-stoppered

                 cylinder previously  rinsed with ether,  add one ml of

                 freshly prepared  10% KI solution.   Shake and let stand

                 one minute.   No yellow color should be observed  in either layer.

          5.7.2  Decompose ether peroxides  by adding 40 g of 30%  ferrous sulfate

                 solution to each   liter of solvent.   CAUTION:  Reaction maybe

                 vigorous if the solvent contains a high concentration of

                 peroxides.

          5.7.3  Distill  deperoxidized ether in glass  and add 2%  ethanol.

-------
                                     1-6




    5.8  Acetonitrile, Heoane, Methanol, Methylene Chloride, Petroleum



         Ether (boiling range 30-60 C) - nanograde, redistill in glass



         if necessary



    5.9  Pesticide Standards - Reference grade.



6.  Calibration :



    6.1  Gas chromatographic operating conditions are considered acceptable



         if the response to dicapthon is at least 50% of full scale when



         < 0.06 ng is injected for electron capture detection and < 100 ng is



         injected for microcoulometric or electrolytic conductivity detection.



         For all quantitative measurements, the detector must be operated



         within its linear response range and the detector noise level should



         be less than 2% of full scale.



    6.2  Standards are injected frequently as a check on the stability of



         operating conditions.  Gas chromatograms of several standard



         pesticides are shown in Figures 1, 2, 3 and 4 and provide reference



         operating conditions for the four recommended columns.



    6.3  The elution order and retention ratios of various organochlorine



         pesticides are provided in Table 1, as a guide.



7-  Quality Control



    7.1  Duplicate and spiked sample analyses are recommended as quality cpntrol



         checks.  When the routine occurrence of a pesticide is being observed,



         the use of quality control charts is recommended (5).



    7.2  Each time a set of samples is extracted, a method blank is determined



         on a volume of distilled water equivalent to that used to dilute the



         sample. •

-------
                                   1-7



8.  Sample Preparation



    8.1  Blend the sample if suspended matter is present and adjust pH to



         near neutral (pH 6.5-7.5) with 50% sulfuric acid or 10 N sodium



         hydroxide.



    8.2  For a sensitivity requirement of 1 yg/1, when using microcoulometric



         or electrolytic conductivity methods for detection, 100 ml or more



         of sample will be required for analysis.  If interferences pose no



         problem, the sensitivity of the electron capture detector should



         permit as little as 50 ml of sample to be used.  Background informa-



         tion on the extent and nature of interferences will assist the



         analyst in choosing the required sample size and preferred detector.



    8.3  Quantitatively transfer the proper aliquot into a two-liter



         separatory funnel and dilute to one liter.



9.  Extraction



    9.1  Add 60 ml of 15% methylene chloride in hexane (v:v)  to the sample



         in the separatory funnel and shake vigorously for two  minutes.



    9.2  Allow the mixed solvent to separate from the sample, then draw the



         water into a one-liter Erlenmeyer flask.  Pour the organic layer



         into a 100 ml beaker and then pass it through a column containing



         3-4 inches of anhydrous sodium sulfate,  and collect  it in a 500  ml



         K-D flask equipped vith a 10 ml ampul.   Return the water  phase  to



         the separatory funnel.   Rinse the Erlenmeyer flask with a second



         60 ml volume of solvent;  add the solvent to the separatory funnel



         and complete the extraction  procedure a  second time.   Perform a



         third extraction in the same manner.



    9.3  Concentrate  the extract in the  K-D evaporator on a hot water bath.

-------
                                      1-8



     9.4  Analyze by gas chromatography unless a need for cleanup is indicated.



          (See Section 10).



10.   Clean-up and.Separation Procedures



     10.1  Interferences in the form of distinct peaks and/or high background



           in the initial gas chromatographic analysis, as well as the physical



           characteristics of the extract (color, cloudiness, viscosity) and



           background knowledge of the sample will indicate whether clean-up



           is required.  When these interfere with measurement of the pesticides,



           or affect column life or detector sensitivity, proceed as directed



           below.



     10.2  Acetonitrile Partition - This procedure is used to isolate fats and



           oils from the sample extracts.  It should be noted that not all



           pesticides are quantitatively recovered by this procedure.  The



           analyst must be aware of this and demonstrate the efficiency of



           the partitioning for specific pesticides.  Of the pesticides listed



           in Scope (1.2) only mirex is not efficiently recovered.



           10.2.1  Quantitatively transfer the previously concentrated extract



                  , to a 125 ml separatory funnel with enough hexane to bring



                   the final volume to 15 ml.  Extract the sample four times



                   by shaking vigorously for one minute with 30 ml portions



                   of hexane-saturated acetonitrile.                       :



           10.2.2  Combine and transfar the acetonitrile phases to a one-liter



                   separatory funnel and add 650 ml of distilled water and



                   40 ml of saturated sodium chloride solution.  Mix thoroughly



                   for 30-45 seconds.  Extract with two 100 ml portions of

-------
                                   1-9






              hexane by vigorously shaking about 15 seconds.




      10.2.3  Combine the hexane extracts in a one-liter separatory funnel




              and wash with two 100 ml portions of distilled water.  Dis-




              card the water layer and pour the hexane layer through a




              3-4 inch anhydrous sodium sulfate column into a 500 ml K-D




              flask equipped with a 10 ml ampul.  Rinse the separatory




              funnel and column with three 10 ml portions of hexane.




      10.2.4  Concentrate the extracts to 6-10 ml in the K-D evaporator




              in a hot water bath.




      10.2.5  Analyze by gas chromatography unless a need for further




              cleanup is indicated.




10.3  Florisil Column Adsorption Chromatography




      10.3.1  Adjust the sample extract volume to 10 ml.




      10.3.2  Place a charge of activated Florisil (weight determined by



              lauric-acid value, see Appendix I) in a Chromaflex  column.




              After settling the Florisil by tapping the column,  add about




              one-half inch layer of anhydrous granular sodium sulfate to




              the top.




      10.3.3  Pre-elute the column,  after cooling, with 50-60 ml  of




              petroleum ether.   Discard the eluate and just prior to



              exposure of the sulfate layer to air, quantitatively transfer




              the sample extract into the column by decantation and subse-




              quent petroleum ether washings.   Adjust the elution rate to



              about 5 ml per minute and,  separately,  collect  up to three




              eluates in 500 ml K-D flasks equipped with 10 ml ampuls.



              (See Eluate Composition 10.4).

-------
                                  1-10
              I.

              Perfcra: the first s?U:C-. on ». •. :>. *\V :n'. .-•: f<% rthvl other  in

              petroleum ether, and the second elution with 200 ml of 15%

              ethyl ether in petroleum ether.  Perform the third elution

              with 200 ml of 50% ethyl ether - petroleum ether and the

              fourth elution with 200 ml of 100% ethyl ether.

      10.3.4  Concentrate the eluates to 6-10 ml in the K-D evaporator

              in a hot water bath.

      10.3.5  Analyze by gas chromatography.
                                                                        i
10.4  Eluate Composition - By using an equivalent quantity of any batch of

      Florisil as determined by its lauric acid value, the pesticides will

      be separated into the eluates indicated below:

                                6% Eluate

             Aldrin               DDT                       Pentachloro-
             BHC                Heptachlor                   nitrobenzene
             Chlordane          Heptachlor Epoxide          Strobane
             ODD                Lindane                     Toxaphene
             DDE                Methoxychlor                Trifluralin
                                  Mirex                     PCB's

                    15% Eluate                50% Eluate

                  Endosulfan  I              Endosulfan II
                    Endrin                  Captan
                  Dieldrin
                  Dichloran
                  Phthalate esters

      Certain thiophosphate pesticides will occur in  each of the above

      fractions as well as the 100% fraction.  For additional information
              /
      regarding eluate composition, refer to the FDA  Pesticide Analytical
                                                                        c
      Manual  (6).

-------
                                    1-11
11.   Calculation of Results



    11.1  Determine the pesticide concentration by using the absolute calibra-



          tion procedure described below or the relative calibration procedure



          described in Part I, Section 3.4.2. (1).



          (1)    Micrograms/liter = (A)  (B)  (Vt)




                                      (Vj)   (Vs)



                 A = ng standard

                     Standard area



                 B » Sample aliquot area



                 V. = Volume of extract injected (ul)
                  i


                 V  = Volume of total extract (yl)



                 V  = Volume of water extracted (ml)



12.   Reporting Results



    12.1  Report results in micrograms per  liter without correction for



          recovery data.  When duplicate and spiked samples are analyzed,all



          data obtained should be reported.

-------
                                        1-12

                                        REFERENCES

 1.   "Method  for Organic  Pesticides  in Water and Wastewater,"  Environmental
     Protection Agency, National  Environmental  Research Center, Cincinnati, Ohio
     45268, 1971.

 2.   Monsanto Methodology for Aroclors - Analysis of Environmental  Materials  for
     Biphenyls, Analytical Chemistry Method 71-35,  Monsanto Company,  St.  Louis,
     Missouri 63166,  1970.

 3.   "Method  for Polychlorinated  Biphenyls  in  Industrial Effluents,"  Environmental
     Protection Agency, National  Environmental  Research Center, Cincinnati, Ohio
     45268, 1973.

 4.   "Method  for Organophosphorus Pesticides in Industrial Effluents,"  Environ-
     mental Protection Agency, National Environmental Research Center, Cincinnati
     Ohio  45268, 1973..

 5.   "Handbook  for Analytical Quality Control  in Water and Wastewater Laboratories,"
     Chapter  6, Section 6.4,  U.S. Environmental Protection Agency,  National Environ-
     mental Research  Center,  Analytical Quality Control Laboratory, Cincinnati,
     Ohio  4S268, 1973.

 6.   "Pesticide Analytical Manual,"  U.S. Dept.  of Health, Education and Welfare,
     Fqod  and Drug Administration, Washington,  D,C.

 7.   "Analysis  of  Pesticide Residues in Human  and Environmental Samples," U.S.
     Environmental Protection Agency, Perrine  Primate Research Laboratories,
     Perrine, Florida  33157, 1971.

 8.   Mills, P.A.,  "Variation of Florisil Activity:   Simple Method for Measuring
     Adsorbent  Capacity and its Use  in Standardizing Florisil Columns," Journal
     of the Association of Official  Analytical  Chemists, 51, 29 (1968).

 9.   Goerlitz,  D.F. and Brown, E., "Methods for Analysis of Organic Substances
     in Water," Techniques of Water  Resources  Investigations of the United States
     Geological Survey, Book 5, Chapter A3, U.S. Department of the  Interior,
     Geological Survey, Washington,  D.C. 20402, 1972, pp. 24-40.
               i
10.   Steere,  N.V., editor, "Handbook of Laboratory  Safety," Chemical  Rubber
     Company, 18901 Cranwood Parkway, Cleveland, Ohio 44128, 1971,  pp. 250-254.

-------
                                       1-15





                                      Table  1




   RETENTION RATIOS OF VARIOUS ORGANOCHLORINE  PESTICIDES  RELATIVE  TO ALDRIN
Liquid Phase

Column Temp.
Argon/Methane
Carrier Flow
Pesticide
Trifluralin
«-BHC
PCNB
Lindane
Dichloran
Heptachlor
Aldrin
Heptachlor Epoxide
Endosulfan I
p,p'-DDE
Dieldrin
Cap tan
Endrin
o,p'-DDT
p,p'-DDD
Endosulfan II
p,p'-DDT
Mi rex
Methoxychlor
Aldrin
(Min absolute)
1.5% OV-17
+
1.95% QF-1
200 C
60 ml/min
RR
0.39
0.54
0.68
0.69
0.77
0.82
1.00
1.54
1.95
2.23
2.40
2.59
2.93
3.16
3.48
3.59
4.18
6.1
7.6

3.5
5%
OV-210
180 C
70 ml/min
RR
1.11
0.64
0.85
0.81
1.29
0.87
1.00
1.93
2.48
2.10
3.00
4.09
3.56
2.70
3.75
4.59
4.07
3.78
6.5

2.6
3%
OV-1 ,
180 C
70 ml/min
RR
0.33
0.35
0.49
0.44
0.49
0.78
1.00
1.28
1.62
2.00
1.93
1.22
2.18
2.69
2.61
2.25
3.50
6.6
5.7

4.0
6% QF-1
+
4% SE-30
200 C
60 ml/min
RR
0.57
0.49
0.63
0.60
0.70
0.83
1.00
1.43
1.79
1.82
2.12
1.94
2.42
2.39
2.55
2.72
3.12
4.79
4.60

5.6
All columns glass, 180 cm I 4 mm ID, solid support Gas-Chrom Q (100/120 mesh)

-------
                                       1-1





                                    •\rn-\PT\  \





13.   Standardization of Florisil Column by Weight Adjustment Bas^d on Ad




     o£ Laurie Acid.




     13.1  A rapid method for determining adsorptive capacity of  Florisil  is




           based on adsorption of lauric acid from hexane solution (6)  (8).




           An excess of lauric acid is used  and amount not adsorbed  is  measured




           by alkali titration.  Weight of lauric acid adsorbed is used to



           calculate, by simple proportion,  equivalent quantities of Florisil




           for batches having different adsorptive capacities.




     13.2  Apparatus




           13.2.1  Buret. -- 25 ml with 1/10 ml graduations.




           13.2.2  Erlenmeyer flasks.  — 125 ml narrow mouth and  25  ml, glass




                   stoppered.




           13.2.3  Pipet. -- 10 and 20 ml transfer.




           13.2.4  Volumetric flasks.  -- 500 ml.




     13.3  Reagents and Solvents
             t


           13.3.1  Alcohol,  ethyl.  —  USP or absolute, neutralized to




                   phenolphthalein.



           13.3.2  Hexane.  — Distilled from all glass apparatus.




           15.3.3  Lauric acid. —Purified,  CP.




           13.5.4  Lauric acid solution. --  Transfer 10.000 g lauric acid  to



                   500 ml volumetric flask,  dissolve in  hexane, and  dilute to




                   500 ml (1 ml = 20 mg).




           13.3.5  Pfafinolphthalein  Indicator. -- Dissolve 1 g in  alcohol and




                   dilute to 100 ml.

-------
                               1-2
      13.3.6  Sodium hydroxide. -- Dissolve 20 g NaOH (pellets, reagent



              grade) in water and dilute to 500 ml (IN).   Dilute 25 ml


              1N_ NaOH to 500 ml with water (0.05N).  Standardize as follows:


              Weigh 100-200 mg lauric acid into 125 ml Erlenmeyer flask.


              Add 50 ml neutralized ethyl alcohol and 3  drops phenol-


              phthalein indicator; titrate to permanent  end point.  Calculate


              mg lauric acid/ml 0.05 N_ NaOH (about 10 mg/ml).
            •

13.4  Procedure


      13.4.1  Transfer 2.000 g Florisil to 25 ml glass stoppered Erlenmeyer


              flasks.   Cover loosely with aluminum foil  and heat overnight


              at 130°C.  Stopper, cool to room temperature, add 20.0 ml


              lauric acid solution (400 mg),  stopper, and shake occasionally


              for 15 min.   Let adsorbent settle and pipet 10.0 ml of


              supernatant into 125 ml Erlenmeyer flask.   Avoid inclusion


              of any Florisil.


      13.4.2  Add 50 ml neutral alcohol and 3 drops indicator solution;


              titrate with 0.05N_ to a permanent end point.


15.5  Calculation of Lauric Acid Value and Adjustment of Column Weight


      13.5.1  Calculate amount of lauric acid adsorbed on Florisil as


              follows:


              Lauric Acid value = mg lauric acid/g Florisil = 200 - (ml


              required for titration X mg lauric acid/ml  0.05N_NaOH).


      13.5.2  To obtain an equivalent quantity of any batch of Florisil,


              divide 110 by lauric acid value for that batch and multiply


              by 20 g.   Veri ;"v proper elution of pesticides by 13.6.

-------
                               1-3
13.6  Test for Proper Elution Pattern and Recovery of Pesticides:



      Prepare a test mixture containing aldrin, heptachlor epoxide,



      p,p'-DDE, dieldrin, Parathion and malathion.  Dieldrin and



      Parathion should elute in the 15% eluate; all but a trace of



      malathion in the 50% eluate and the others in the 6% eluate.

-------
                                           T
        25
20
 15                10
RETENTION TIME IN MINUTES
Figure  1. Column Packing: 1.5% OV-17 + 1.95% QF-1, Carrier  Gas:  Argon/Methane at 60 ml/min,
                     Column Temperature: 200 C,  Detector: Electron  Capture.

-------
15
                          10                 5                 0
                      RETENTION TIME IN MINUTES
Figure  2. Column Packing:  5%  OV-210,  Carrier Gas: Argon/Methans
          at 70  ml/min,  Column Temperature: 180 C, Detector:
          Electron Capture.

-------
                         I	I	I
                                                                            I   I   I
   25
20
 15                10
RETENTION TIME IN MINUTES
Figure  3. Column Packing: 6% QF-1 + 4%  SE-30, Carrier  Gas: Argon/Methane  at 60 ml/min,
                   Column  Temperature: 200  C,  Detector: Electron  Capture.

-------
3 •
S
3
           25
20
5
                               15                10
                            RETENTION TIME IN MINUTES
Figure  4.  Column Packing:  3%  OV-1, Carrier  Gas:  Argon/Methane  at  70 ml/min,
          Column Temperature:  180 C, Detector: Electron Capture.

-------
                          NO.  7
METHODS FOR CHEMICAL ANALYSIS OF WATER AND WASTES

-------
         CE NO.  8
DETERMINING SELENIUM

-------
                           Reprinted from Atomic Absorption Newsletter, Volume 14, No. 5
                 DETERMINING SELENIUM  IN WATER, WASTEWATER,  SEDIMENT, AND
                     SLUDGE BY FLAMELESS ATOMIC ABSORPTION SPECTROSCOPY
                                    Theodore  D. Martin*  and John F. Kopp
                                       Environmental Protection Agency
                               Environmental Monitoring & Support Laboratory
                                            Cincinnati, Ohio 45268
                                                      and
                                               Richard D. Ediger
                                         The Perkin-Elmer Corporation
                                            Lombard, Illinois 60148
 ABSTRACT
 A method  has been  developed for the
 determination of selenium in freshwater,
 wastewater, sediment,  and sludge by
 flamelese  atomic  absorption  spectros-
 copy. A simple and rapid sample prep-
 aration is given with a description of the
 interferences that  affect  the  analysis.
 Nickel nitrate is added to both standards
 and samples to prevent losses by volatil-
 ization. The method demonstrates good
 precision  with day-to-day  variation of
 the absorbance values at the 0.25 and 2.5
 nanogram level (25 nl of 10 and 100 ng
 Se/1) varying only ±11.6% and ±4.1%
 respectively, at the 95% confidence level.
 The sensitivity of the  method is 20 pico-
 grams which for many tap, surface, and
 well waters extends the detection limit
 to 0.2 ug Se/1 without the use of scale
 expansion.
 INTRODUCTION
RESUME
On a develop pe  une methode pour la
determination du selenium dans des eaux
fraiches et residuelles, des sediments et
des boues par spectroscopie d'absorp-
tion  atomique  sans  flarame. On  donne
une methode simple et  rapide pour la
preparation   des  echantillons   ainsi
qu'une description des interferences qui
affectent 1'analyse.
On fait un ajout de nitrate de nickel aux
standards et aux echantillons de maniere
a empecher des pertes par volatilisation.
La methode  offre une bonne precision
avec  une variation des valeurs d'absorb-
ance de jour en  jour de ± 11,6% et
± 4,1% seulement au seuil de confiance
de 95% et ce respectivement pour des
valeurs de 0,25 et 2,5 nanogrammes (25
jil de 10 et 100 ng  Se/1). La sensibilite de
la methode est de 20 picogrammes, ce
qui pour de nombreuses eaux de  distri-
bution, de surface et  de puits donne une
detection limite de  0,2 ng  Se/1, sans
usage d'expansion d'echelle.
 ZUSAMMENFA5SUNG
 Eine Methode zur Bestimmung von Selen
 in Frischwasser, Abwasser, Sedimenten
 und Klarschlamm wurde mittels  der
 Sammenlosen  Atomabsorptions - Spek-
 troskopie entwickelt.  Es wird auf eine
 einfache und rasche Probenaufbereitung
 hingewiesen und die, die Analyse beein-
 flussenden Interferenzen angegeben.
 Nickelnitrat wird  sowohl zu  den Stan-
 dards alsauch zu den Proben zugesetzt,
 um eventuelle Verluste  durch Fliichtig-
 keit zu  vermeiden. Die  Methode zeigt
 fiir Extinktionswerte beim 0,25 und 24
 ng-Niveau (25 nl von  10 und  100 ng
 Se/1) eine gute Prazision und Reprodu-
zierbarkeit,  mit  Abweichungen von nur
 ±11,6%  beziehungsweise ±4,1%,  bei
einer 95%igen Sicherheit. Die Empfind-
lichkeit der Methode von 20 pg erweitert
die Nachweisgrenze bis  0,2 ng Se/1  fiir
viele  Leitungs.-0berflachen.-und  Quel-
lenwasser, ohne der Anwendung  einer
Skalendehnung.
   The analytical determination of selenium has long been
a problem to the analytical chemist. It is similar to arsenic
in toxicity and reactivity yet is probably much more diffi-
cult to detect and measure. The procedure given in Stan-
dard Methods for the Examination of Water and  Waste-
water, 13th Ed., 1971, is time-consuming, subject to many
interferences, and relatively insensitive thus requiring  a
large volume of sample.  Therefore, it is often omitted from
routine analysis. Additionally,  the colorimetric  reagent
often used I diaminobenzidine) has been placed on the pos-
sible carcinogenic listing.
   The selenium concentration of most  finished waters is
less than 10 ug, 1. However, the use of selenium in industry
is growing. A major use  of selenium is  in the glass industry
to  color glass a deep red and to neutralize iron color. Sele-
nium is known to be present in almost all types of paper.
Selenium may be present in soils both as selenite and sele-
nate. Thus, it is likely to be found in surface waters. Al-
though trace  amounts of selenium have been shown to be
nutritionally  beneficial in  some animal diets, exposure to
higher concentrations produces toxic effects. There are also
some implications that selenium is a carcinogen.
  In addition to the four valence states in which selenium
may exist,  a variety  of organo-selenium compounds is
known. Therefore, to ensure measurement of total sele-
nium, any method de%ased must include an oxidation step.
During this  digestion,  it is  most important to maintain
oxidizing conditions. Inorganic selenium is  not appreci-
* Author to whom correspondence should be addressed.
                    ably volatilized during digestion in a mixture of nitric and
                    perchloric acids (1, 2, 3, 4), except in the presence of such
                    a large excess of organic material that charring occurs (3,
                    4). In general, selenium may be lost  from acid selenite
                    solutions by reducing but not by oxidizing agents.
                      In attempting to avoid volatilization losses, several meth-
                    ods effected dissolution of the sample through combustion
                    with oxygen in closed systems  (1, 5, 6, 7). This can be done
                    either with a  Schoniger  combustion  flask or in a Parr
                    bomb. Watkinson (8), after comparing wet oxidation with
                    the oxygen flask combustion, found no significant differ-
                    ence between the results of the two methods. He preferred
                    oxidation with nitric and perchloric  acids.  The  use  of
                    perchloric acid, however, is discouraged because of safety
                    reasons.
                      Since most natural waters  and waste effluents contain
                    low concentrations of selenium, conventional  atomic ab-
                    sorption has not been used for the analysis because of its
                    relatively poor sensitivity. Since it forms a hydride similar
                    to arsenic,  several investigators  have  applied  the arsine-
                    type procedure  with subsequent introduction in an argon-
                    hydrogen flame to selenium (9). This technique has the
                    advantage of eliminating interference resulting from the
                    matrix  effect and  improving  the detection limit.  Many
                    problems, however, have been encountered in determining
                    selenium  in domestic and  industrial wastewaters particu-
                    larly the interference of organics, hizh copper concentra-
                    tions, and difficulty in forming the hydride.
                      With the advent of flameless atomization devices and
                    electrodelesa discharge lamps the analytical working ran^e
                    for many  elements has been extended.
ATOMIC  ABSORPTION  NEWSLETTER
Vol. 14, No. 5. September-October 1975
             AA-861
                                109

-------
  Baird  et al.  (101.  recently reported a  flameless  AAS
method for the determination of selenium in wastewater
employing a carbon rod analyzer. The obvious advantage
of this mode of analysis  is the absence of the usual high
levels of flame background normally responsible  for de-
creased sensitivity. Bairc  did observe the need for a pre-
liminary  digestion with  nitric  and  perchloric  acid  to
oxidize organic material and solubiiize the selenium before
injections into the carbon rod. Replicate analyses of sam-
ples containing approximately 10 (ig of selenium per liter
gave a relative standard deviation of 6.8%. Because of the
safety  factor, the author?  of this paper  prefer a digestion
step using a combination of nitric acid-hydrogen peroxide.
This combination allows the integrity of the sample  to be
broken down and all of the selenium to be solubilized  while
a condition of oxidation  is maintained. Complete ashing
occurs during the charring step after the sample has been
injected into the furnace.
  Because of its volatility, the possible loss of selenium has
been a point of concern in flameless analysis. In the devel-
opment of a furnace method for the analysis of selenium
in biological materials. Ediger (11. 12) determined that
the addition of nickel nitrate prior to the drying step pro-
duces  a  condition  where  high chairing temperatures
(1200° to 1500^ i  can  be tolerated without the  .loss of
selenium. This condition  facilitates complete ashing and
the removal of some matrix constituents which may  cause
subsequent interference during the  atomization. Welcher.
et al. (13 I have also demonstrated the stability of selenium
in the presence of nickel or other heavy metals in the deter-
mination  of  trace elements  in high  temperature alloys.
Recently. Henn  (14'i demonstrated  the same effect with
the use of molybdenum after separation  of the selenium
from metallic  interferences with a cation exchange resin.
  This paper describes a  method incorporating the nitric
acid-hydrogen peroxide digestion procedure, followed  by
the addition of nickel nitrate and the use of the HGA-2000
Graphite Furnace in connection with a  selenium EDL for
the determination of selenium in water, wastewater, sedi-
ments, and sludges  while focusing on  the problems en-
countered durins the analvsis.
EQUIPMENT

   A Perkin-Elmer Model 503 atomic absorption spectro-
photometer equipped with a  Perkin-Elmer Model HGA-
2000 Graphite Furnace,  a  Deuterium  Background Cor-
rector,  and a  Perkin-Elmer  selenium  Electrodeless Dis-
charge Lamp  (EDL)  was used for the analysis. The spec-
trophotometer  was operated  in the peak  read  mode.  A
Perkin-Elmer Model 056 recorder on 10 millivolt span was
used to record the absorbance signals. The Deuterium Back-
ground Corrector was used to compensate for non-specific
absorption using nitrogen at a flow rate of 2 liters.'min  to
purge the optics.  The selenium EDL  was  operated at 9
watts, with a wavelength setting of 196.0 nm and a spectral
slit width of 0.7 nm. All equipment requiring 120 volts was
operated  on regulated voltage with the  spectropbotometer
connected to a Stabiline saturable reactor autotransformer
voltage regulator to insure'stable power.
   The HGA-2000 Graphite Furnace was programmed for
drying at 125QC (.with varying times depending on  the vol-
ume of aliquot injected): 30-sec charring at 1500°C*:
and 10-sec atomization at 2700CC. Argon was used as the
furnace purge gas at a flow rate of 3 divisions, and the flow
was interrupted automatically during atomization.
  Drying times of 20 sec were used for volumes of 5 and 10
ul: 30 sec for 25 ul; 50 sec for 50 ul: 65 sec for 75 ul: and
80 sec for 100 ul.
  Eppendorf microliter pipets  with disposable tips were
used to inject the samples into the furnace.

REAGENTS AND STANDARDS
Nitric Acid
  (HN03), concentrated, ACS reagent grade, redistilled.
Hydrogen Peroxide
  (H202), 307c, ACS reagent grade.
Standard Selenium Solution
  A stock solution of 1000 mg Se/1 was  prepared  by dis-
solving 0.3453 grams of selenous acid (actual assay 9^.6%
HoSeOa) in 200 ml of deionized distilled water. Diluie
working standards  (1, 2, 5,10,40, 50 and  100 pig Se/1 were
prepared from a diluted stock solution of 10 mg Se/1 by
withdrawing the appropriate aliquot, adding to it 1 ml of
cone. HNO?, 2 ml  30% H202 and diluting to  100 ml with
deionized distilled water.
Nickel Nitrate
   (1% Ni solution) — Dissolve 4.956 g of ACS reagent
grade Ni(N03)2-6H20 in 100 ml  of deionized distilled
water.
Nickel Nitrate
  (5% Ni Solution)  — Dissolve 24.780 g of ACS reagent
grade Ni(N03)2-6 H20 in 100 ml of deionized distilled
water.

SAMPLE PREPARATION AND PROCEDURE

  Detailed procedures on sample preparation and the final
concentration of nickel depend on sample type, matrix, and
concentration of selenium to be determined.  In all cases
where total selenium  is to  be determined, the sample is
subjected to vigorous oxidation to solubilize the selenium.

Well and Surface Water
  Transfer 100 ml  of well-mixed sample to a 250-ml Griffin
beaker,  add 3 ml  cone, redistilled  HNO-j and 5 ml 30%
H=02. Heat  for  one hr at 95 °C or until the volume is
slightly less than 50 ml. Cool and bring back to 50 ml with
deionized distilled  water. Pipet 5 ml of this digested solu-
tion  into a 10-ml  volumetric flask, add  1 ml of the 1%
nickel nitrate solution  and dilute to 10 ml with deionized
distilled water. The sample is now ready for analysis.
  Since  the  nickel  concentration  is  0.1%,   the  sample
should be compared to a standard curve constructed from
standards also  containing 0.1% nickel.  The  aliquot size
used for injection  into the furnace should be the same for
both  samples  and  standards.  Recommended  volume  for
this type of sample  is 25 to 100 ul.
  To verify the absence of matrix or chemical interference,
an aliquot of the digest solution should be spiked with a
known  amount  of selenium,  nickel  nitrate  added, and
analyzed in the same manner. The actual signal compared
to the expected response will indicate the presence of any
significant interference. Those samples where interference
is detected must either have the interference  reduced by
dilution or be analyzed by the method of standard addi-
tions. (See discussion on interferences.)
  Many surface water samples having low dissolved solids
(400 mg/'l) may be concentrated 5X during the digestion
step.  Even though this technique  extends the detection
'Since there are differences between individual furnaces  and the
 reading and setting  of the maximum allowable charring tempera-
 ture, each furnace should be checked to determine  the maximum
 charring temperature before beginning the analyses.
110
               ATOMIC  ABSORPTION NEWSLETTER
               VoL 14,  No. 5, September-October 1975

-------
 limit, the solution must still be checked for interference by
 spiking an aliquot of the concentrate and performing the
 analysis.
   Samples  with  sulfate  concentration  higher  than 200
 mg/1 should be analyzed in the presence of 1% Ni. Samples
 are prepared and analyzed as previously described except
 the addition of nickel  nitrate should be 2.0 ml of the 5%
 Ni solution diluted to 10 ml. Results should be determined
 from a standard curve prepared from standards containing
 1% Ni.

 Industrial Waste Effluent
   Sample should be prepared  in the same manner as sur-
 face water, but the nickel concentration in the final dilution
 should be 1%. Results of many industrial effluents can be
 determined from a standard curve, but again each must be
 checked for possible interferences before this assumption
 can  be made. In some cases sample dilution may be re-
 quired. A typical set of data from Se standards in  a 1%
 Ni matrix is listed in Table I. If it is necessary to use the
 method of standard additions, the size of the aliquot used
 for injection into the furnace will depend on  the reproduc-
 ibility of signal response and the amount of interference
 encountered. An aliquot of 25 ul was used for the work on
 this type of sample as reported in this paper.

 Sediments and Sludges
   Weigh and transfer to a 250-ml Griffin beaker a 0.5-g
 portion of a sample which has been dried at 60°C, pul-
 verized, and thoroughly mixed. Add 5 ml of cone. HNOj
 and cover with a watch glass. The sample should be heated
 at 95°C and refluxed to near dryness. Allow  the sample to
 cool, add another 5  ml  of  cone. HN03 and repeat the
 digestion step. After the second reflux step has been com-
 pleted, allow the sample to cooL add 3 ml of cone. HN03
 and 10 ml 30% H202. Return the beaker to the hot plate
 for wanning to start vigorous reaction. When the reaction
 has commenced, immediately remove the beaker from hot
 plate. After  effervescence has subsided, return the covered
 beaker to the hot plate and reflux for 15 minutes.  After
 cooling, dilute the sample to 50 ml with deionized distilled
 water. Mix and withdraw a  5-ml aliquot, to be diluted to
 10 mL for analysis by the method of standard additions.
 Each final solution should contain 1% Ni (2.0 ml of 5%
 Ni solution)  and suspended solids should be permitted to
 settle before analysis.  It  is suggested that solutions  used
 for analysis by the  method of standard additions contain
 5 ml of sample plus 15, 30,  and 45 ug Se/1. Because the
 possibility of encountering severe interferences is greatest
 in this type of sample, a 5-ul aliquot should be  used for
 furnace injection. A detection limit of 5 }ig/g sample can
 be achieved with this method using a 5-jjil injection.

 ANAIYT1CAL PROCEDURE

   The instrument should be operated using the conditions
 as listed in the section  on Equipment. As previously men-
 tioned, the size of the  aliquot used  for furnace injection
 will depend on the sample type as well as the matrix. When
 the method  of standard  additions is required, a  linear
curve over the entire range of the additions is necessary for
 the results to be considered valid.
  The life and performance of the furnace tube using this
method will mainly be affected by the number of analyses
completed. Many  tubes have lasted  for  more than  100
firings but it  is recommended, because of varying sample
types, that the tube be replaced after 100 firings. Prolonged
 use of a given tube will result in elevated values, sometime
 exceeding the expected value by more than 10%; but since
 the increase is a gradual drift,  there is no loss of precision
 in consecutive analyses. For those samples which  have a
 complex matrix including metals of low volatility and  re-
 quiring the 1% nickel matrix, a conditioning cleaning burn
 after each analysis may prove helpful. This can be accom-
 plished  by eliminating the drying and charring step, and
 atomizing at 2700°C for 15 sec without gas interrupt

 RESULTS AND DISCUSSION

 Effect of Nickel
   The addition  of nickel to the sample  serves three pur-
 poses. First, it is believed to form a stable selenide com-
 pound at the beginning of the char cycle thereby reducing
 the volatility  of  selenium. This  allows the use of elevated
 charring temperature for complete ashing and  volatiliza-
 tion of some possible interfering and non-specific absorb-
 ing substances.  The second advantage of the nickel is the
 increase in sensitivity gained because of the enhancement
 effect. To demonstrate this effect a new graphite tube was
 placed in the furnace, and the charring temperature was
 set at 200°C. Average absorbances of 0.110 and 0.225  re-
 spectively  were  recorded for a 25-ul injection of a 40  ug
 Se/1 solution first without, and then with the addition of
 nickel (1000 mg/1). The absorbance value (0.225) for the
 charring temperature  of 200° C is, for all practical pur-
 poses, the same as when the charring temperature for the
 same selenium solution is raised to 1500°C (Abs=0.235i.
 This comparison, with and without the  nickel, indicates
 about a  two-fold enhancement because of the nickel. This
 enhancement  may be due to a decrease in the rate  cf
 atomization or a change in the efficiency of the atomization.
 Under the standard conditions given, a nickel concentration
 of 100 mg/1 to 2000 mg/1 gives a similar enhancement
 but when the nickel concentration is increased to 10,000
 mg/1, or 1%, the absorbance for 25 ul  of a 40 ug  Se/1
 solution drops to 0.170. Since the amount of nickel in the
 furnace  during  atomization  is critical to the  signal re-
 sponse, it must be controlled and the same quantity  must
 be present for both standards and samples.
   Thirdly, the nickel serves as a stable matrix for those tap.
 surface and well waters which have low concentrations  of
 trace metals  and sulfate ion. thereby permitting the use
 of a standard curve of the same nickel concentration.

 Standard Curve
   Table 1 shows the average  absorbance and  relative
 standard deviation values for a composite standard curve
 in a 0.1%  Ni matrix over a  concentration range of 5 to
 100 [ig Se/1. The volume of the aliquot used for the injec-
 tion for each standard was 25 ul. These standard data are
 the result of values collected on 9 different days over a
 period of 3 weeks. The composite data reflect normal dailv
 variations  in  instrumental parameters and the  effect of
 different graphite tubes. Selenium  is linear up  to an ab-
 sorbance value of 0.4 in a 0.1% nickel matrix. A working
 detection limit using this technique is 2 (ig Se/1. This
 detection limit can be extended to 1 ug Se/1 using a 100-ul
 aliquot injection or to as little as 0.2 ug Se/1 if the sample
 is  first concentrated five times by evaporation, and a 50-ul
 aliquot used for the injection. In both cases the concentra-
 tion  of the constituents in the sample matrix will be the
determining factor. To verify  this procedure and standard
data, quality control check samples supplied by the Quality
Assurance Branch of the Environmental  Monitoring and
ATOMIC  ABSORPTION NEWSLETTER
Vol. 14, No. 5,  September-October 1975
                                                   111

-------
                       TABLE I
      Selenium Standard Data in 0.1 % Ni Matrix'
Se Concenrrafion
ug//iter
5
10
20
40
50
100
Average
Afasorbonce
0.035
0.065
0.118
0.235
0.290
0.540
% Relative
Std Deviation +
14.2
11.6
9.3
7.2
6.4
4.1
    the 95% confidence level.

      Selenium Standard Data in 1 % Ni Matrix*
Se Concentration
ug//iter
10
20
40
50
100
Average
Afaiorbance
0.046
0.091
0.170
0.219
0.413
*25-u,l sample aiiquots.

Support Laboratory in Cincinnati were analyzed at the 4,
16, and  48 ug/l levels with recoveries of 90%, 97%, and
96% respectively.

Water Matrix — Sulfate Interference
   A  major concern  of  any  analytical  technique  is  the
possible effect of the common minerals and anions present
and their concentration  on the analytical result.  Table II
lists a variety  of parameters,  their concentrations,  and
the selenium response observed. Examination  of the data
indicates an inverse relationship between the concentration
of magnesium  and sulfate and the selenium absorbance.
Since it is known that a concentration of magnesium as
high as 200 mg/1 has no effect on the selenium response,
the increase in  the suppression of selenium is attributed to
the increasing  concentration of sulfate. Table HI  is also
evidence of sulfate interference.  Section  (A) shows the
effect of large concentrations of sulfate. Section (B) shows
the effect in more detail over a small concentration range.
Section (C) shows that the  degree of the  sulfate suppres-
sion can be reduced by increasing the amount of nickel
present during the analysis. If  the  concentration of the
nickel in the injection aliquot is increased to 1%  (10,000
mg/1), an injection of 50 ug of sulfate (50 ul of 1000 mg
SOt/l) will only cause  a  15% suppression to the signal
generated by 1  nanogram of Se (25 ul of 40 |ig Se/1). See
Table V for a comparison.

Chloride and Nitrate Interference
   Chloride and nitrate also  affect selenium absorption. In
both the 0.1% and the 1% nickel matrix, chloride concen-
trations greater than 800 mg/1 cause a significant suppres-
sion (greater than 5%)  of the absorbance. If the chloride
is  increased to 2000 mg/1, the suppression  in 0.1% Ni and
1% Ni is approximately 15% and 30%, respectively. Thus
an increase  in nickel concentration does not decrease the
suppressive effect of chloride, and therefore the method as
described is not applicable to the analysis  of seawater
and brines.
   In  selenium  solutions containing 1% v/v cone. HNQs
there is an 80% reduction in the Se absorbance when the
nickel is omitted, but in a 0.1% Ni  matrix with. 3% v/v
cone. HN03 no  reduction was observed. At  levels above 3%
nitrate,  interference is  encountered  in the 0.1%  nickel
                                                     TABLE II
                        Selenium Absorbance in Six Synthetic Surface Water Matrix Solutions
                                            of Various Concentrations*
Element, Anion
or Measured
Parameter
Calcium
Magnesium
Sodium
Potassium
Alkalinity
Chloride
Total hardness
Total dissolved solids
Sulfats
Volume of Aliquot
25 ul
% response
50 ul
% response
lOOul
% response
Distilled
Water •
Solution
0
0
0
0
0
0
0
0
0

0.122
100%
0.224
100%
0.345
100%
Concentration mg/liter
I
90
21
82
16
180
174
310
570
84

—
—
0.346
100%
2
180
25
210
32
280
350
550
1200
260
Se
—
0.223
100%
0.321
93%
3
180
41
260
32
280
350
630
1450
440
4
180
41
390
32
560
350
600
1760
440
5
360
41
390
63
560
700
1200
2300
440
6
360
82
520
63
560
700
1240
2900
870
Absorbanca Values
0.104
85%
0.181
81%
0.269
78%
0.100
82%
0.189
84%
0.263
76%
0.109
89%
0.191
85%
0.279
81%
0.097
80%
0.164
73%
0.237
69%
                      'Each of the 6 synthetic matrix solutions and the distilled water solution contained 20
                       u.g Se/1 in 0.1% Ni. 1% v/v cone. HNO*. 2^0 v/v 30% H30S.
112
               ATOMIC ABSORPTION NEWSLETTER
               Voi.  14, No.  5, September-October 1975

-------
                                                      TABU III
                                      Effect of Suifata on Selenium Absorbance
Volume of
Suifate
Aliquot
0
25 |il
(A) 25 |il
25 ul
25 ul


0
25 ul
50 uJ
25 ul
(B) 50 ul
100 |il
50 ul
100 |il


0 (it
25 |il
50 jil
(C) 50 uJ
50 ul
75 ul
Concentration
0
500
1000
2000
4500


0
270
180
450
270
180
450
270


0
450
270
360
450
360
Total ug
SO4//n/ection
0
13
25
50
113


0
6.8
9.0
11.3
13.5
18.0
22.5
27.0


0
11.3
13.5
18.0
22.5
27.0
Se
Absorbance*
0.30
0.22
0.16
0.10
0.07
Se
Absorbancef
0.230
0.230
0.215
0.199
0.166
0.140
0.126
0.107
Se
Absorbance +
0.230
0.212
0.201
0.187
0.168
0.142
% Suppress/on of
Se Absorbance
—
27%
45%
66%
76%


—
0
6%
13%
28%
39%
45%
53%


—
8%
13%
19%
27%
38%
                  *Se absorbance value and corresponding suppression is the result of a 25-ul injection of SO ng Se/1
                   in 0.1% Ni (25 ug iYi/irt;ec«on;, 1% v/v cone. HNOj 2% v/v 30% HiO» with the listed quantity
                   of suifate pipetted on top of the Se injection.
                  tSame conditions as  in * except 25 ul of 40 |ig Se/1 in 0.1% Ni f2$ pg  Ni/injection), 1% v/v
                   cone. HNOs. 2% v/v 30% H»05 was used for injection.
                  "Same conditions as in * except 50  (il of 20 |ig Se/1 in 0.1% Ni (SO \ig  Hi/injection), 1% v/v
                   cone. HN03,2% v/v 30% H»0j was  used for injection.
matrix. Although this interference can  be somewhat re-
duced and stabilized by the use of a longer charring cycle,
concentrations of over 30,000 mg NOa/l should be avoided.

Single Metal Interference
   Table IV lists concentrations of single  metal solutions
and the degree to which these metals  affect the Se absorb-
ance in 0.1% Ni  matrix.  These approximate  results  are
given  as an indication of when  the analyst can no longer
reliably use a standard curve prepared in 0.1% Ni for the
determination. Special  attention should be  given to  the
concentration of Fe, Sn, Si, Al,  Mn, V, and Cr. Although
seldom present at  these concentrations in surface  and  tap
water, there may be other types  of environmental  samples
including sludges and sediments  where these elements may
exist in even greater concentrations than listed. It has been
determined that increasing the concentration of the nickel
to 1% decreases the suppressive effect of many metals. A
comparison of the suppressive effect of some of the more
critical metals in 0.1%.and \%  nickel matrix is  given in
Table V. Although the selenium response is lower in 1% Ni
than in 0.1 ^ Ni when other metals are absent, the same is
not necessarily true with the addition of these metals ,is evi-
dent in Table V. This  phenomenon can be an  advantage in
eliminating large suppressive effects when  analyzing sam-
ples with a complex matrix.
   Since all of the metals tested have a concentration which
can be tolerated without causing an interference, the deter-
mination of their composite effect at those concentrations
both  with  and  without the  synthetic  water  matrix was
important. Table VI lists the matrix parameters and trace
metals, their concentrations, the affected selenium absorb-
ance, and percent suppression in both 0,1% and 1% nickel
solutions. In reviewing Table  VI, it is apparent that there
is a composite effect in 0.1% Ni but not in 1% Ni and that
the combination of matrix and  trace metals produces in-
creased suppression in 0.1% Ni. This suppressive effect is
strong evidence for using 1% nickel when analyzing sam-
ples with a complex matrix or ones that contain ions or
trace metals at concentrations  known to interfere.  The
sample may be analyzed using either a standard curve
prepared in 1% Ni or, if necessary,  by the technique of the
method of  standard additions.  Whenever  possible,  and
especially for tap water and  clean, low  dissolved solids
surface water, the 0.1% nickel matrix should  be  used be-
cause of the added sensitivity.

Dissolved and Suspended So/ids
   In  considering the  effect of  dissolved and suspended
solids, it was determined that the nature or chemical com-
position of  the solids rather than the physical state was the
important factor. Also the amount  of  an  interfering  sub-
ATOMIC ABSORPTION NEWSLETTER
Vol. 14, No. 5, September-October 1975
                                                    113

-------
                                   TABLE IV
Suppression Effects on
Element
Ag
Al
As
3
Bo
Be
Cd
Co
Cr
Cu
Fe
U
Mo
Mn
Ni
P
Pb
Sb
Si
Sn
Sr
Ti
Tl
V
In
Concentration
W/TIC/I HOJ
No Effect,
mg/l
400
20
40
300
400
10
400
400
50
100
4
300
200
20
300
100
200
40
10
2
400
200
40
20
400
Selenium Absorbance* of Single Metal Solutions +
Concentration
mg/l
_
40
100
400
—
20
—
—
100
200
10
400
400
50
400
200
300
50
20
4
—
400
200
30
—
Suppress/on

10%
10%
10%

5%


10%
10%
10%
10%
10%
10%
10%
25%
15%
10%
10%
10%

5%
20%
10%

Concentration
mg/l
—
200
400
—
—
40
—
—
200
—
20
—
—
200
—
400
400
100
40
10
—
—
400
100
—
Suppress/on

65%
50%


20%


55%

30%


50%

75%
30%
20%
50%
40%


30%
30%

•A 25-M.l injection of 40 n« Se/1 in 0.1% Ni, 1% v/v cone. HNOj, 2% v/v 30% H,0» was used for this
 comparative work.
'The approximate suppression values are the result of a 25-nl injection of the listed concentrations
 being pipetted on top of the selenium injection.
                                    TABLE V
            Comparison of Suppression Effects of Trace Metal Solutions
            on Selenium Absorbance in 0.1% Ni and 1% Ni Solutions
Trace Metal Solution
                             25u/40ugSe//
                                ;n0.7%N/


50 ul
50 ul
50 ul
50 ul
50 ul
50 ul
50ul
50 ul

_
lOOmgAl/l
100mgCr/l
1QOmgCu/l
lOOmgFe/1
lOOmgSi/l
100 mg Sn/l
100mgV/l
1000mgSO4/l
Absorbance
0.235
0.075
0.095
0.195
0.120
0.025
0.015
0.085
0.047
% Suppression
—
68
60
17
49
89
94
64
80
Absorbance
0.170
0.105
0.140
0.154
0.126
0.086
0.103
0.148
0.145
% Suppression
_
38
18
9
26
49
39
13
15
                                                         ATOMIC ABSORPTION NEWSLETTER
                                                         Vol.  14, No.  5, September-October  1975

-------
                                                       TABLE VI
                        Comparison of Selenium Absorbance in 0.1 % Ni and 1 % Ni Solutions
                                    Containing Synthetic Matrix and Trace Metals
                                          Matrix
      TVaco Metals












Solution
Content
Se
Se + matrix
Se + metals
Se + matrix
•4* metals
Element
Calcium
Magnesium
Sodium
Potassium
Alkalinity
Chloride
Total hardness
Total dissolved
solids
Sulfate

Aliquot
Injection
Volume
25 ul
25 uJ
25 uJ

25 uJ
Cone.
180 mg/l
20 mg/l
190 mg/l
32 mg/l
280 mg/l
350 mg/l
600 mg/l

1160 mg/l
220 mg/l

0.7%
Element
Sn
Be
Fe
Si
V
Mo
Al
As
Cu
Mn
P
N;
Se C30 \ig/l) % Enhanc. or
Absorbance Suppression
0.170
0.175
0.150

0.106
—
+ 3%
-12%

-38%
Cone.
1 mg/l
2 mg/l
2 mg/l
5 mg/l
5 mg/l
10 mg/l
20 mg/l
20 mg/l
20 mg/l
20 mg/l
20 mg/l
7%Ni
Se (24 ug/U % En/ianc. or
Absorbanca Suppression
0.105 -
0.101 -4%
0.111 +6%

0.109 +4%
 stance present during atomization is the important consid-
 eration — not whether injected as a dissolved or suspended
 solid. Review of Table II reveals  that a change in  total
 dissolved solids in  the synthetic  water matrix from  1450
 mg/l to 2300 mg/l did not produce  a significant difference
 in  the  selenium absorbance.  Since  the  total dissolved
 solids for most surface water are below 2000 mg/l, solids
 should not be a problem in the analysis  of water samples
 provided that the non-specific absorption does not exceed
 the background correction capability of the instrument.

 Sample Analysis and Recovery Data
   Analytical  results and spike recovery  on  a variety of
 sample types are given in Table VII. The sample prepara-
 tion used was that described in this paper. The results were
 determined  from standard curves prepared in 0.1%  and
 1%  nickel solutions, and by utilizing the method of stan-
 dard additions.

 CONCLUSION

   This method utilizing the HGA-2000 Graphite  Furnace
 provides a rapid procedure for analyzing a variety  of water
 samples  for  selenium. After the sample  preparation  and
 solubilization.  the samples are diluted in either 0.1% or
 1% nickel matrix and compared to a standard curve of the
 same matri* to determine the  result of the  analysis.  The
 method demonstrates  satisfactory precision  and  is  suffi-
ciently sensitive with  a  working detection limit  of 2 ug
 Se 1 which can be extend
-------
                                                      TABLE VII
                             Results of Sample Analyses and Recovery Data for Selenium
No. of
Sample Type Somp/es
Anoryzed
Tap water
Surface water


Well water
(N. Mexico)



Well water
(N. Mexico)


Drinking water
for animal ex-
posure studies
Sewage plant
effluent
Industrial
waste effluent

Landfill
leachate
Ocean dis-
posal


Sludges
Sediments
Solid gelatin
ref. std.

2
2


6




4



4


2

6
3

1
1
1



2
3
1


No. of
PosmVe
Occur-
rences
0
0


1




0



4


0

0
0

0
1
0



0
0
1


Technique
Amount of
Detected Deter-
mination
N.D. < 2 ug/l
N.D. <2 ug/l


5ug/I




N.D. <0-5 ug/l



57 ug/l
58 ug/l
o?M/!
N.D. <2ug/I

N.D. < 5 ug/l
N.D. < 4 ug/l

N.D. < 10 ug/l
50 ug/l
N.D. 
-------