5838
                  Method 524.  Measurement of  Purgeable  Organic
                           Compounds  In Drinking Water
                     by Gas Chromatography/Mass Spectrometry
                                  February  1983
                                Ann Alford-Stevens
                               Games W»  Eichelberger
                                 William L. Budde
                       Physical and Chemical Methods Branch
                  Environmental  Monitoring and Support Laboratory
                        Office of Research and Development
                       U.S. Environmental Protection Agency
                              Cincinnati, Ohio 45268

-------
  OATE
             UNITED STATES ENVIRONMENTAL PROTECTION AGENCY
                        FEB 2 2 1983
SUBJECT  Nationwide  Approval of Alternate Test Procedure
                    01 Trihalomethanes
  FROM vrctor Ox  Kimm, Director
      Office of  Drinking Water (WH-550)

    T0'Regional Administrators
      Listed below is an alternate test procedure for determining
      trihalomethanes by a gas chromatography-mass spectrcrr.etry
      (GC-MS) method which I have approved for nationwide use for
      "National Interim Primary Drinking Water Regulation"  (NIPDWR)
      compliance monitoring.

      This method was developed by the Environmental Monitoring and
      Support Laboratory (EMSL) in Cincinnati for the measurement
      of purgeable organic compounds in drinking water.  Unlike the
      previously approved Method 501.3 which involves selected-ion
      monitoring with a mass spectrometer, this method permits the
      acquisition of complete mass spectra.  In addition, this
      method can be used for the measurement of many other volatile
      organic compounds along with the trihalomethanes.  EMSL's
      data show that the precision and accuracy of this method is
      comparable to that of EPA's approved methods.   (Previously
      approved  methods for trihalomethanes are 501.1, 501.2 and
      501.3.)  Although detection limits for trihalomethanes are
      not  as low as those obtained with other approved methodology,
      they are  entirely adequate for the current maximum contaminant
      level.
     Measurement
     Total  Trihalomethanes
                                 Method

                                 ^Method 524 - "Measurement of
                                  Purgeable Organic Compounds
                                  in urinking water by Gas
                                  Chromatography/Mass
                                  Spectrometry".
j-V	--TT--   r - - -- - _	
z/-.ddi uionul information  on  this  method is available from the
 Envi .vor.rnontal Monitoring and  Support  Laboratory, 26 West St
 £ti;eot, Cincinnati, Ohio 45268.

cc:   Robo.'.-t L. Booth, Acting Director,  EMSL
     Ann Alford-SLevonn, Chomist,  EMSL
     Roqi OD.J 1 Writf?r Supply  Representatives
     H'.'fji onn 1 Quality AG.'jur.irice  Officers
                                                                    Clair

-------
                                     INDEX
Section
Number                      Subject

 1              Scope  and  Application
 2              Summary of Method
 3              Definitions
 4              Interferences
 5              Safety
 6              Apparatus  and  Equipment
 7              Reagents and Consumable Materials
 8              Sample Collection,  Preservation  and  Handling
 9              Calibration
10              Quality Control
11              Procedure
12              Calculations
13              Precision  and  Accuracy
                References
                                     TABLES
1.  Ion Abundance  Criteria for  p-Bromofluorobenzene.
2.  Single Laboratory  Method  Efficiency  Data  for  Purgeable Organic  Compounds
    Measured with  GC/MS.
3.  Acceptable  Storage Times  for  River and  Drinking Water  Samples Containing
    Halogenated Aliphatic Analytes.
4.  Storage Time Data  for River and  Drinking  Water Samples Containing
    Aromatic Analytes.
5.  Single Laboratory  Accuracy  and Precision  Data for  Purgeable Organic
    Compounds Measured with GC/MS.

-------
            Method 524.   Measurement of Purgeable Organic Compounds
           in  Drinking Water By Gas Chromatography/Mass Spectrometry
1.  SCOPE AND APPLICATION

    1.1  This is a general purpose method that provides procedures for
         identification and measurement of purgeable organic compounds  in
         finished drinking water, raw source water, or drinking water in any
         treatment stage.  The method is applicable to a wide range of
         organic compounds that have sufficiently high volatility and low
         water solubility to be removed from water samples with purge and
         trap procedures.  Single-laboratory method efficiency, accuracy and
         precision data have been determined for the following compounds:
                                    Chemical Abstracts Service^
         Analyte
         benzene
         bis(2-chloroisopropyl) ether
         bromoch1orometh ane
         bromodichloromethane
         4-bromof1uoroben zene
         bromoform
         carbon tetrachloride
         chlorobenzene
         chlorod ibromomethane
         chloroform
         chloromethane
         1,2-dibromo-3-chloropropane
         1,2-dichlorobenzene
         1,3-dichlorobenzene
         1,2-dichloroethane
         1,1-dichloroethene
         trans-1,2-dichloroethene
         cis-1,3-dichloropropene
         methylene chloride
         styrene (ethenylbenzene)
         1,1,2,2-tetrachloroethane
         tetrachloroethene
         toluene
         1,1,1-trichloroethane
         1,1,2-tr1chloroethane
         trichloroethene
         vinyl chloride
         p-xylene
Registry Number (CASRN)
      71-43-2
   39638-32-9
      74-97-5
      75-27-4
     460-00-4
      75-25-2
      56-23-5
     108-90-7
     124-48-1
      67-66-3
      74-87-3
      96-12-8
      95-50-1
     541-73-1
     107-06-2
      75-35-4
     156-60-5
   10061-01-5
      75-09-2
     TOO-42-5
      79-34-5
     127-18-4
     108-88-3
      71-55-6
      79-00-5
      79-01-6
      75-01-4
     106-42-3
STORET
Number
 34030
 34283
 77297
 32101

 32104
 32102
 34301
 32105
 32106
 34418

 34536
 34566
 34531
 34501
 34546
 34699
 34423
 77128
 34516
 34475
 34010
 34506
 34511
 39180
 30175
         A laboratory may use this method to detect and measure additional
         analytes after the laboratory obtains acceptable (defined in
         Section 10) accuracy and precision data for each added analyte.

-------
    1.2  Detection Limits (MDLs) (1) are compound dependent,  varying  with
         purging efficiency and concentration, where MDL  is defined as  the
         statistically calculated minimum amount that can be  measured with
         99% confidence that the reported value is greater than  zero.   For
         the listed analytes in reagent water, MDLs vary from 0.07 to 11.2
         ug/L.  The applicable concentration range of this method  is
         compound and instrument dependent but is approximately  0.2 ug  to
         200 ug of analyte per liter of undiluted sample.  Analytes that  are
         inefficiently purged from water will not be detected  when present
         at low concentrations, but they can be measured with  acceptable
         accuracy and precision when present in sufficient amounts.

    1.3  Determination of some individual components of complex  mixtures  may
         be hampered by insufficient chromatographic resolution  and/or  by
         large differences in concentrations of individual components.

2.  SUMMARY OF METHOD

    Highly volatile organic compounds with low water solubility  are removed
    (purged} from the sample matrix by bubbling helium through a 25-mL
    aqueous sample.  Purged sample components are trapped in  a stainless
    steel tube containing suitable sorbent materials.  When purging is
    complete, the sorbent tube is heated and backflushed with  helium to
    desorb purged sample components into a gas chromatograph  (GC) interfaced
    to a mass spectrometer (MS).  Compounds eluting from the  GC  column  are
    tentatively identified by comparing their mass spectra to  reference
    spectra in a data base.  Tentative identifications are confirmed by
    analyzing standards under the same conditions used for samples and
    comparing resultant mass spectra and GC retention times.   Each identi-
    fied component is measured by relating the MS response for an appro-
    priate selected ion produced by that compound to the MS response for  the
    same ion from that same compound in an external standard or  for another
    ion produced by a compound that is used as an internal standard.

3.  DEFINITIONS

    External standard — a known amount of a pure compound that  is analyzed
    with the same procedures and conditions that are used to analyze samples
    containing that compound.  From measured detector responses  to known
    amounts of the external standard, a concentration of that same compound
    can be calculated from measured detector response to that compound  in a
    sample analyzed with the same procedures.

    Internal standard — a pure compound added to a sample in known amounts
    and used to calibrate concentration measurements of other compounds that
    are sample components.  The internal standard must be a compound that is
    not a sample component.

    Field duplicates — two samples taken at the same time and place under
    identical circumstances and treated exactly the same throughout field
    and laboratory procedures.  Analysis of field duplicates  indicates the
    precision associated with sample collection, preservation  and storage,
    as well as with laboratory procedures.

-------
Field reagent blank — reagent water placed  in  a  sample  container  in  the
laboratory and treated as a sample in all respects,  including exposure
to sampling, site conditions, storage, preservation and all  analytical
procedures.

Laboratory control standard — a solution of analytes prepared  in  the
laboratory by dissolving known amounts of pure  compounds in  a known
amount of reagent water.  In this method, the laboratory control
standard is prepared by adding appropriate volumes of the secondary
dilution standard solution and the internal  standard/surrogate  compound
spiking solution to reagent water.

Laboratory duplicates — two aliquots of the same sample that are
treated exactly the same throughout laboratory  analytical procedures.
Analysis of laboratory duplicates indicates  precision associated with
laboratory procedures but not with sample collection, preservation or
storage procedures.

Laboratory reagent blank — a 25-mL portion  of  reagent water placed in
the purging apparatus and analyzed as if it  were  a sample.

Performance evaluation sample — a methanol  solution of  method  analytes
distributed by the Quality Assurance Branch  (QA8), Environmental
Monitoring and Support Laboratory, USEPA, Cincinnati, Ohio, to  multiple
laboratories for analysis.  A small volume of the methanol  solution is
added to a known volume of reagent water and analyzed with  procedures
used for samples.  Results of analyses are used by the QAB  to determine
statistically the accuracy and precision that can be expected when a
method is performed by competent analysts.   Analyte  true values are
unknown to the analyst.

Quality control check sample — a methanol solution  containing  known
concentrations of analytes prepared by a laboratory  other than  the
laboratory performing the analysis.  The analyzing laboratory uses this
solution to demonstrate that it can obtain acceptable identifications
and measurements with a method.  A small volume of the methanol solution
is added to a known volume of reagent water  and analyzed with procedures
used for samples.  True values of analytes are  known by  the  analyst.

Secondary dilution standard — a methanol solution of analytes  prepared
in the laboratory from stock standard solutions and  diluted  as  needed to
prepare aqueous calibration solutions and laboratory control standards.

Stock standard solution — a concentrated solution containing a
certified standard that is a method analyte, or a concentrated  methanol
solution of an analyte prepared in the laboratory with an assayed
reference compound.  Stock standard solutions are used to prepare
secondary standard solutions.

Surrogate compound ~ a compound that is not expected to be  found  in  the
sample, is added to the original environmental  sample to monitor perform-
ance, and is measured with the same procedures  used  to measure  sample
components.

-------
4.  INTERFERENCES

    4.1  Samples may be  contaminated  during  shipment  or  storage  by diffusion
         of volatile organics  through the  sample  bottle  septum seal.   Field
         reagent blanks  must be  analyzed to  determine when  sampling and
         storage procedures have not  prevented  contamination.

    4.2  During analysis, major  contaminant  sources are  volatile materials
         in the laboratory and impurities  in the  inert purging gas and in
         the sorbent trap.  Analyses  of field reagent blanks and laboratory
         reagent blanks  provide  information  about the presence of contami-
         nants.

    4.3  Interfering contamination  may occur when a sample  containing  low
         concentrations  of volatile organic  compounds is analyzed
         immediately after a sample containing  relatively high concentra-
         tions of  volatile organic  compounds.   A  preventive technique  is
         between-sample  rinsing  of  the purging  apparatus and sample syringes
         with two  portions of  reagent water. After analysis of  a sample
         containing high concentrations of volatile organic compounds, the
         system should be baked  for 10 min by passing helium through the
         sample purging  chamber  into  the heated (180°) sorbent trap.   One
         or more laboratory reagent blanks should be  analyzed to ensure that
         accurate  values are obtained for  the next sample.

5.  SAFETY

    5.1  The toxicity or carcinogenicity of  chemicals used  in this method
         has not been precisely  defined; each chemical should be treated as
         a potential health hazard, arid exposure  to these chemicals should
         be minimized.   Each laboratory is responsible for  maintaining
         awareness of OSHA regulations regarding  safe handling of chemicals
         used in this method.  Additional  references  to  laboratory safety
         are cited (2-4).

    5.2  The following method  analytes have  been  tentatively classified as
         known or  suspected human or  mammalian  carcinogens: benzene*  carbon
         tetrachloride,  chloroform, and vinyl chloride.   Pure standard
         materials and stock standard solutions of these compounds should be
         handled in a hood.

6.  APPARATUS AND  EQUIPMENT

    6.1  Sample containers —  120-mL  or larger  glass  bottles each equipped
         with a screw cap and  a  polytetrafluoroethylene-faced silicone
         septum.
    6.2  ftjrge and trap  device consisting  of sample purging chamber, sorbent
         trap and  desorber.  (Acceptable devices  are  commercially available.)

         6.2.1  The all  glass  sample  purging chamber  holds  25-mL samples
                with < 15 ml of  gaseous headspace between the water column
                and the  trap.  The  helium  purge gas passes  through the water
                column as finely divided bubbles  (optimum diameter of  <3 mm

-------
               at the origin).  The purge  gas  must  be  introduced at a
               point <5 mm from the base of  the  water  column.

     6.2.2  The stainless steel sorbent trap is  25  cm  long  by  2.5 mm ID
            and is packed with 8 cm of Tenax-GCR, 8 cm of  silica gel,
            and 8 cm of charcoal, in that  order  with respect to  the
            inlet end of the trap.  The charcoal  is not necessary for
            listed analytes but is necessary if  fluorine-substituted
            methanes and ethanes (fluorocarbons)  are among  additional
            analytes.  When analytes do not  include fluorocarbons, the
            charcoal may be eliminated, and  the  amount of Tenax-GCR
            may be increased.  A trap with different dimensions  can be
            used if it has been evaluated  and  found to perform satisfac-
            torily (i.e., provides method  efficiencies equal to  or
            better than those in Table 2).   Before  initial  use,  the trap
            should be conditioned overnight  at 18QOC by backflushing
            with helium flow of at least 20  mL/min.  Each day  the trap
            should be conditioned for 10 min at  180°C  with  back-
            flushing.

     6.2.3  The desorber should be capable of  rapidly  heating  the trap
            to 180°C.  The trap section containing  Tenax-6CR should
            not be heated to higher than 180°C,  and the temperature of
            the other sections should not  exceed  200°C.

6.3  SYRINGES AND SYRINGE VALVES

     6.3.1  Two 25-mL glass hypodermic syringes with Luer-LokR tip (if
            applicable to the purging device being  used).

     6.3.2  One 5-mL gas-tight syringe with  shutoff valve.

     6.3.3  Two two-way syringe valves with  Luer  ends  (if applicable to
            the purging device being used).

     6.3.4  Micro syringes, various sizes.

6.4  MISCELLANEOUS

     6.4.1  Standard solution storage containers  — 10-mL bottles with
            polytetrafluoroethylene-lined  screw  caps.

     6.4.2  Analytical balance capable of  weighing  0.0001 g accurately.

     6.4.3  Helium purge gas, as contaminant free as possible.

6.5  GAS CHROMATOGRAPH/MASS SPECTROMETER/DATA  SYSTEM (GC/MS/DS)

     6.5.1  The 6C must be capable of temperature programming. Any
            column (either packed or capillary) that provides  data with
            adequate accuracy and precision  (Sect.  10)  can  be  used.  If
            a packed column is used, the GC  usually is interfaced to the
            MS with an all-glass enrichment  device  and an all-glass

-------
                transfer line, but any enrichment device or transfer  line
                can be used if performance specifications described in this
                method can be demonstrated with it.  If a capillary column
                is used, an enrichment device is not needed.  A recommended
                packed GC column for the listed analytes is 1.8 m  long by 2
                mm ID glass packed with 1% SP-1000 on 60/80 mesh Carbopack
                B.  Recommended operating parameters for that column  are:
                helium carrier gas flow rate of 30 mL/min and temperature of
                45°C for 4 min, increased to 230°C at a rate of
                8°C/min, and isothermal at 230°C for at least 25 min  or
                until all expected analytes elute.  An alternative recom-
                mended packed column is 1.8 m long by 2 mm ID glass or
                stainless steel packed with 0.2% Carbowax 1500 on 80/100
                mesh Carbopack C.

         6.5.2  Mass spectral data are obtained with electron-impact
                ionization at a nominal electron energy of 70 eV.  The mass
                spectrometer must be capable of scanning from 35 to 450 amu
                every 7 s or less and must produce a mass spectrum that
                meets all criteria in Table 1 when 50 ng or less of
                £-bromofluorobenzene (BFB) is introduced into the GC.  To
                ensure sufficient precision of mass spectral data, the
                desirable MS scan rate allows acquisition of at least five
                spectra while a sample component elutes from the GC.  With
                capillary columns which produce narrower peaks than packed
                columns that criterion may not be feasible and adequate
                precision with fewer spectra per GC peak must be
                demonstrated (Sect. 10).

         6.5.3  An interfaced data system (DS) is required to acquire,
                store, reduce and output mass spectral data.  The computer
                software must allow searching any GC/MS data file for ions
                of a specific mass and plotting ion abundances versus time
                or scan number.  This type of plot is defined as an
                extracted ion current profile (EICP).  Software must  also
                allow integrating the abundance in any EICP between speci-
                fied time or scan number limits.

7.   REAGENTS AND CONSUMABLE MATERIALS

    7.1  SORBENT TRAP PACKING MATERIALS

         7.1.1  Polymer based on 2,6-diphenyl-£-phenylene oxide — 60/80
                mesh Tenax-6CR, chromatographic grade, or equivalent.

         7.1.2  Coconut charcoal — 26 mesh.

         7.1.3  Silica gel — 35/60 mesh, Davison Chemical grade 15,  or
                equivalent.

    7.2  REAGENTS

         7.2.1  Methanol — pesticide quality or equivalent.

-------
     7.2.2  Reagent water — water  in which an  interferent  is  not
            observed at the method  detection  limit of  the compound  of
            interest.  Prepare reagent water by.  passing tap  water through
            a filter bed containing about 0.5 kg of  activated  carbon,  by
            using a water purification system,  or by boiling distilled
            water for  15 min followed by a 1 h  purge with inert gas
            while the  water temperature is held at 90°C.  Store in
            clean, narrow-mouth bottles with polytetrafluoroethylene-
            lined septa and screw caps.

     7.2.3  Sodium thiosulfate or sodium sulfite —  granular,  ACS
            reagent grade.

7.3  STOCK STANDARD SOLUTIONS — These solutions may be purchased as
     certified solutions or prepared from pure  standard materials using
     the following procedures:

     7.3.1  Place about 9.8 ml of methanol in a 10-ml  ground-glass
            stoppered  volumetric flask.  Allow  the flask to stand
            unstoppered for about 10 min or until all  alcohol-wetted
            surfaces have dried.  Weigh the flask to the nearest 0.1
            mg.

     7.3.2  If the analyte is a liquid at room  temperature, with a
            100-uL syringe immediately add two  or more drops of assayed
            reference  compound to the flask.  (The liquid must fall
            directly into the alcohol without contacting the flask).   If
            the analyte 1s a gas at room temperature,  fill  a 5-mL valved
            gas-tight  syringe with  the standard to the 5.0-mL  mark,
        .    lower the  needle to 5 mm above the  methanol meniscus, and
            slowly inject the standard into the neck of the flask.  (The
            gas will rapidly dissolve in the methanol.)

     7.3.3  Reweigh the flask, dilute to volume, stopper, and  mix by
            inverting  several times.

     7.3.4  From the net weight gain, calculate the  concentration in
            micrograms per micro liter.  When assayed compound  purity  is
            >9656, the  unconnected weight may be used to calculate
            concentration.

     7.3.5  Store stock standard solutions with minimal headspace in
            polytetraf1uoroethylene-1ined screw-capped bottles.
            Methanol solutions of listed liquid analytes are stable for
            at least four weeks when stored at  4°C.  Methanol
            solutions  prepared from listed gaseous analytes are not
            stable for more than one week when  stored  at <0°C; at room
            temperature, they must  be discarded after  one day.

7.4  SECONDARY DILUTION STANDARD — Use stock standard solutions to
     prepare a secondary dilution standard solution  that contains the
     analytes in methanol.  The secondary, dilution standard should  be
     prepared at a concentration that can be easily  diluted to prepare

-------
         aqueous calibration solutions (Section 9.2.1.2) at concentrations
         that will bracket the working concentration range.  Store the
         secondary dilution standard solution with minimal headspace and
         check frequently for signs of deterioration or evaporation,
         especially just before preparing calibration solutions from it.

    7.5  INTERNAL STANDARD SPIKING SOLUTION AND SURROGATE COMPOUND SPIKING
         SOLUTION — Prepare a solution of fluorobenzene in methanol at a
         concentration that allows use of 2 to 10 uL to add an appropriate
         amount of fluorobenzene to each sample; this amount should be
         approximately the same as the amount of the analyte to be measured.
         If the internal standard technique is used, fluorobenzene serves as
         the internal standard.  If the external standard technique is used,
         fluorobenzene is a surrogate compound added to each sample to
         monitor method performance.  Fluorobenzene was selected because it
         is stable in aqueous solutions, is. efficiently purged, does not
         occur naturally, and is not commercially produced in bulk
         quantities but is available as a laboratory reagent chemical.

8.  SAMPLE COLLECTION. PRESERVATION AND HANDLING

    8.1  Collect all samples in duplicate.  Fill sample bottles to over-
         flowing.  No air bubbles should pass through the sample as the
         bottle is filled, or be trapped in the sample when the bottle is
         sealed.  Keep samples sealed from collection time until analysis.
         Maximum storage times vary with analytes of concern.  Recent
         studies (5-6) provided data indicating appropriate storage times
         for samples (river and drinking water) containing compounds that
         are potential method analytes (Tables 3 and 4).

         8.1.1  When sampling from a water tap, open the tap and allow the
                system to flush until the water temperature has stabilized
                (usually about 10 min).  Adjust the flow to about 500 mL/min
                and collect duplicate samples from the flowing stream.

         8.1.2  When sampling from an open body of water, fill a 1-qt
                wide-mouth bottle with sample from a representative area,
                and carefully fill duplicate sample bottles from the 1-qt
                bottle.

    8.2  SAMPLE PRESERVATION

         8.2.1  If styrene (which reacts with chlorine) and/or trihalo-
                methanes are to be measured in a sample expected to contain
                residual chlorine, add a reducing agent, sodium thiosulfate
                or sodium sulfite (30 mg per 120-mL sample for up to 5 ppm
                chlorine) to the empty sample bottle before it is shipped to
                the sampling site.

                NOTE:  Some possible analytes may be unstable in the
                presence of reducing agent.  Data (5) indicate that sodium
                sulfite should not be used if analytes include
                chloromethane, 1,1-dichloroethene, 1,1-dichloropropene,

-------
            2-chloroethyl ethyl ether, or 1,1,2,2-tetrachloroethane;
            sodium thiosulfate is not recommended if analytes include
            chloromethane or 1,2-dibromoethane.

     8.2.2  Much remains to be learned about biological degradation of
            aromatic hydrocarbon analytes.  Currently, two preservation
            techniques are recommended but both have negative aspects.

            8.2.2.1 Hydrochloric acid may be used at the sampling site
                    to adjust the sample pH to < 2; the major disadvan-
                    tages of this procedure are that shipping acid is
                    restricted by federal regulations and that effects
                    of low pH on other analytes (such as organohalides)
                    are largely unknown.

            8.2.2.2 Mercuric chloride may be added to the sample bottle
                    in amounts to produce a concentration of 10 mg/L.
                    This may be added to the sample at the sampling site
                    or to the sample bottle in the laboratory before
                    shipping to the sampling site.  A major disadvantage
                    of mercuric chloride is that it is a highly toxic
                    chemical; it must be handled with caution, and
                    samples containing it must be disposed with appro-
                    priate procedures.

            8.2.2.3 If analytes include both aromatic hydrocarbons and
                    styrene or trihalomethanes, current recommendations
                    are either to add both preservative types (reducing
                    agent along with acid or mercuric chloride) or to
                    collect two samples with the appropriate preserva-
                    tive type in each.

     8.2.3  After addition of preservative(s), seal the sample bottle
            and shake vigorously for 1 min.

8.3  FIELD BLANKS

     8.3.1  Duplicate field reagent blanks must be handled along with
            each sample set, which is composed of the samples collected
            from the same general sample site at approximately the same
            time.  At the laboratory, fill field blank sample bottles
            with reagent water, seal, and ship to the sampling site
            along with empty sample bottles and back to the laboratory
            with filled sample bottles.  Wherever a set of samples is
            shipped and stored, it is accompanied by appropriate blanks.

     8.3.2  When reducing agent or preservative(s) is added to samples,
            use the same procedures used for samples to add the same
            amount to blanks.  The reducing agent can be added in the
            laboratory.

-------
9.  CALIBRATION

    9.1  INITIAL CALIBRATION

         9.1.1  CALIBRATION SOLUTIONS
                9.1.1.1 At  least  three  calibration  solutions,  each  contain-
                        ing  a  standard  of  each  analyte,  are needed.   (More
                        than three  calibration  solutions may be required if
                        analytes  are  numerous.)  One  calibration solution
                        should contain  each  analyte at  a concentration
                        approaching but greater than  the method detection
                        limit  (Table  5)  for  that compound;  the other two
                        solutions should contain analytes at concentrations
                        that bracket  the range  expected  in  samples.   For
                        example,  if the detection limit  for a particular
                        analyte is  0.2  ug/L,  and a  25-mL sample expected to
                        contain approximately 5 ug/L  is  analyzed, aqueous
                        solutions of  standards  should be prepared at
                        concentrations  of  0.3 yg/L, 5 ug/L, and 10  yg/L.

                9.1.1.2 To  prepare  calibration  solutions, add  appropriate
                        volumes (uL)  of the  secondary dilution standard
                        solution  to aliquots  of reagent  water.  Remove the
                        plunger from  a  25-mL  syringe  and attach a closed
                        syringe valve.   Fill  the syringe with reagent water,
                        replace the plunger,  and compress the water. Open
                        the  syringe valve  and vent  air.   Adjust the  reagent
                        water  volume  to 25 mL and add a  carefully measured
                        aliquot of  2.0  to  18.0  uL of  the secondary dilution
                        standard  through the  valve  bore. Add  the appro-
                        priate amount (> 2.0  uL) of the  internal standard
                        surrogate spiking  solution  through  the valve bore,
                        but  do not  add  more  than 20 yL total volume  of
                        methanol  solution.

                        NOTE:   If appropriate concentrations cannot  be
                        prepared  without adding more  than 20 yL of the
                        secondary dilution standard to  25 mL of reagent
                        water, prepare  a new  secondary dilution standard.
                        If  less than  2.0 yL  must be added to obtain  appro-
                        priate concentrations,  prepare  a larger volume of
                        the  calibration solution in a volumetric flask.   Mix
                        by  inverting  the flask  several  times,  and transfer a
                        25-mL  portion into the  sample syringe.  The  remain-
                        ing  solution  may be  stored  in screwcap vials with no
                        headspace.   If  aromatic compounds are among calibra-
                        tion solution components, do  not store for  more  than

-------
                    1 h.  Other aqueous  solutions  can  be  stored  for up
                    to 24 h.

     9.1.2  Analyze .triplicate aliquots  of  each  calibration  solution
            with the procedures to be used  to  analyze  samples.

            9.1.2.1 If the external  standard technique  is  being  used,
                    prepare a concentration calibration curve for  each
                    analyte by plotting  integrated  abundances of the ion
                    characteristic of that  compound as  a  function  of the
                    concentration.   If the  ratio of ion abundance  to
                    amount of analyte is constant  (< 10%  relative
                    standard deviation)  throughout  the  concentration
                    range, the average ratio may be used  instead of a
                    calibration curve.

            9.1.2.2 If the internal  standard technique  is  being  used,
                    calculate the mass spectrometer response to  each
                    compound relative to fluorobenzene, the  internal
                    standard.  Calculate the response  factor (RF)  with
                    the equation,
                        RF
-
            where   Ax  *    integrated  abundance  of  the  selected  ion  for
                            the  analyte standard;
                    As  *    integrated  abundance  of  the  selected  ion  for
                            the  internal  standard;
                    Qs  •    quantity  of internal  standard;  and
                    Qx  *    quantity  of analyte standard.

            RF is a unitless  number; units used  to  express quantities  of
            analyte and internal  standard must be equivalent.   Ideally,
            the response factor for  each analyte should be independent
            of analyte  quantity for  the  working  range of the
            calibration, but  required  linearity  will vary  with  required
            accuracy of analyte concentration measurements.  Generally,
            acceptable  variations of mean RF values are ±  15% over a
            concentration range of two orders of magnitude and  ± 10% RSD
            of values obtained  from  analyses of  triplicate aliquots  of
            each concentration  calibration solution.  For  an analyte
            with non-linear RF, a calibration curve of  Areax/Areas
            plotted versus  Qx way be used to determine  an  analyte
            concentration.

9.2  DAILY CALIBRATION  — Check calibration data each day  by measurement
     of one or more laboratory  control standards or calibration  solu-
     tions.  If the expected  ion  abundance was observed (Sect.  10.5) for
     50 ng of the MS performance  standard but the absolute ion  abundance

-------
         measured for any analyte varies from expected abundance by more
         than 15%, prepare and analyze a fresh calibration solution to
         determine if the problem is being caused by deterioration of the
         calibration solution or by a malfunction in the purge and trap
         apparatus.  When the internal standard technique is being used,
         verify each day that response factors have not changed.  When
         changes occur (> 10% relative standard deviation), prepare and
         analyze new standard solutions to determine new response factors.

         NOTE:  Some analysts have observed marked deterioration of MS
         response after the initial purge and trap analysis each day; if
         this phenomenon is observed, perform one purge/desorb cycle before
         checking MS performance and calibration data.

10. QUALITY CONTROL

    10.1 Minimum quality control requirements consist of:

         10.1.1 initial demonstration of laboratory analytical capability
                (efficiency, accuracy and precision procedures, Sect. 10),

         10.1.2 analysis of an MS performance standard and a laboratory
                control standard near the beginning of each 8-h work period,

         10.1.3 analysis of a field reagent blank along with each sample set,

         10.1.4 analysis of a laboratory reagent blank when the field
                reagent blank contains analytes at concentrations above the
                method detection limits,

         10.1.5 quarterly analysis of a quality control check sample, (if
                available for analytes of concern), and

         10.1.6 continued maintenance of performance records to define the
                quality of generated data.

    10.2 METHOD EFFICIENCY — For each analyte, calculate method efficiency
         by comparing the detector response when the compound is introduced
         by syringe injection with the detector response when the same
         amount is introduced by purging, trapping, and desorption.  Because
         of the calibration technique used in this method, high efficiency
         is not required for acceptable precision and accuracy, but low
         method efficiency may cause unacceptably high detection limits.
         Measure method efficiency for each analyte whenever the analytical
         system undergoes major modification, such as replacement of trap
         packing.

         10.2.1 Analyze at least five laboratory control standards with the
                purge, trap, desorption and 6C/MS detection procedures.
                Interspersed among these five or more analyses, inject two
                or more aliquots of the secondary dilution standard solution

-------
            (Section 7.4) directly into the GC to  introduce each  analyte
            in an amount equal to that introduced  by purge and  trap
            procedures.  Use the same MS data acquisition parameters  for
            injected analytes as those used for purge and trap  proce-
            dures.

     10.2.2 Calculate the method efficiency (E) for each analyte  in each
            aliquot of the laboratory control standard with the equation:


                          E . A-   •  100   ,
                               Ai

            where Ap =   ion abundance of compound introduced with purge
                         and trap techniques, and
                  A.J *   ion abundance produced by an equal amount of
                         the same compound when injected.

            For this calculation, use data obtained from an injection
            either closely preceding or following  the purge and trap
            analysis from which data are used.

     10.2.3 Calculate the mean method efficiency for each analyte.
            Acceptable detection limits usually can be achieved even  if
            the mean method efficiency is only 20  to 30%.

10.3 ACCURACY — To determine accuracy, analyze duplicate aliquots of a
     quality control (QC) check sample containing  known amounts of
     analytes of concern.  QC check  samples for some, but not all listed
     analytes, currently are available from the U.S. Environmental
     Protection Agency, Environmental Monitoring and Support Laboratory,
     Quality Assurance Branch, Cincinnati, Ohio  45268; alternatively
     certified standard solutions may be purchased from commercial
     vendors.

     10.3.1 When using the external  standard procedure, calculate
            accuracy as the ion abundance found in the QC sample
            solution expressed as a  percentage (P) of the ion abundance
            found in the external standard solution:

                                Ax
                        P =    -T£—  . 100   ,
            where Ax -   abundance of  ion used to measure  an  analyte  in
                         an aliquot of the QC check  sample, and
                  As =   abundance of  ion used to measure  an  equal
                         amount of the same analyte  treated as an
                         external standard.

     10.3.2 When using the internal standard procedure,  fluorobenzene in
            the solution of analyte standards is the internal standard.

-------
            Calculate response factors (Sect. 9.3.4) for each analyte
            relative to fluorobenzene. With these response factors,
            calculate accuracy with data acquired for the QC check
            sample:

                      P =   Ax  .  100
A                                    DP
                             e>  •   l»r

            where Ax » abundance of ion used to measure an analyte in
                       an aliquot of the QC check sample,
                  As - abundance of ion used to measure
                       fluorobenzene in the same aliquot, and
                  RF « response factor of the particular
                       analyte relative to fluorobenzene.

               NOTE:  The internal standard concentration is constant in
               calibration solutions and all samples for which the
               calibration solutions are used (Section 9.3).

     10.3.3 For each analyte, the mean accuracy should be in the range
            of 85 to 115X.  For some listed analytes, this may not be
            feasible for low concentration measurements.

10.4 PRECISION

     10.4.1 For each analyte, calculate method precision as the standard
            deviation (s, in yg/L) of the replicate measured values
            obtained in the accuracy calculations:

                                      n             n
                                             9           /%
                           .  _    n j .   X^  -  4_iX-i £
                                            n (n-1)

            where n = number of measurements for each analyte, and
                  X = individual measured value.

     10.4.2 For the set of measured values for each analyte, calculate
            the dispersion as the percent relative standard deviation
            (RSO):

                          RSD =   s   .  100
                                  C

            where s « standard deviation, and
                  C s mean observed concentration.

     10.4.3 Adequate precision is obtained when the relative standard
            deviation is £20%.  For some listed analytes, this may not
            be feasible foir low concentration measurements.

-------
10.5 MS PERFORMANCE STANDARD

     10.5.1 Near the beginning of each 8-h work period  in which analytes
            are to be measured, measure the mass spectrum produced by  50
            ng of p_-bromofluorobenzene (BFB) to ensure  that  it meets
            performance criteria (Table 1).  BFB may be introduced into
            the MS either by syringe  injection or through the purge and
            trap system.  It may be a component of the  laboratory
            control standard analyzed daily to check calibration (Sect.
            7).  Measure the entire mass spectrum at an MS scan rate
            that produces at least five spectra for the BFB  GC peak but
            does not exceed 7 s per spectrum.  Although acquisition of
            five spectra per BFB GC peak may not be feasible when capil-
            lary columns are.used, BFB performance criteria  still must
            be met.  If the BFB spectrum is unacceptable, adjust GC/MS
            operating parameters until an acceptable spectrum is pro-
            duced before samples are  analyzed.

     10.5.2 Record the absolute ion abundance detected  for 50 ng of
            BFB.  If ion abundance varies more than ±  10% from the
            expected number, check the GC/MS system to  locate and
            correct the problem.   Preparation of a new  calibration curve
            may be necessary if the system is operating acceptably but
            with decreased sensitivity.

10.6 LABORATORY CONTROL STANDARD — To demonstrate that the  calibration •
     curve is still valid, analyze a  laboratory control standard at the
     beginning of each 8-h work period.

     10.6.1 For each analyte to be measured, select a concentration
            representative of its occurrence in drinking water samples.

     10.6.2 Prepare the laboratory control standard with either of the
            following procedures:

            10.6.2.1  From stock standard solutions, prepare a
                      laboratory control standard concentrate in
                      methanol.  This solution should contain analytes
                      at concentrations 2500 times those selected as
                      representative  concentrations.  Add 10 yL of the
                      laboratory control standard concentrate to a 25-mL
                      aliquot of reagent water.

            10.6.2.2  Add 2 to 18 uL  of the secondary dilution standard
                      to 25 mL of reagent water contained in the sample
                      syringe.

     10.6.3 Add an appropriate volume of the internal  standard/surrogate
            spiking solution and analyze with the same  procedures (Sect.
            11) to be used for samples.

-------
      10.6.4 Determine calibration acceptability and appropriate remedial
             actions, if needed.  (For the external standard technique,
             see Sect. .9.1.2.1; for the internal standard technique, see
             Sect. 9.1.2.2.)

10.7  MONITORING THE SURROGATE COMPOUND/INTERNAL STANDARD - Because all
      samples and laboratory control standards contain equal amounts of
      the internal standard/surrogate compound, use the absolute ion
      abundance for the characteristic ion of that compound, fluoro--
      benzene, to monitor system performance.  If for any sample, the
      absolute ion abundance varies more than 15% from that observed in
      the previous sample or laboratory control standard, do not report
      analyte values obtained for that sample, and take remedial actions
      to solve the system performance problem.

10.8  FIELD REAGENT BLANKS ~ Analyze a field reagent blank along with
      each sample set.  If a field reagent blank contains analytes at
      concentrations above the method detection limits, analyze a
      laboratory reagent blank.  If one or more analytes that are not
      detected at concentrations above method detection limits in the
      laboratory reagent blank are detected in significant amounts in the
      field blank, sampling or storage procedures have not prevented
      sample contamination, and the appropriate analyte measurement(s)
      must be discarded.

10.9  At least quarterly, analyze a quality control check sample obtained
      from the U.S. Environmental Protection Agency, Environmental
      Monitoring and Support Laboratory, Quality Assurance Branch,
      Cincinnati, Ohio.  Quality control check samples currently are
      available for some but not all listed analytes.  If measured
      analyte concentrations are not within ±20% of true values, check
      the entire analytical procedure to locate and correct the problem
      source.

10.10 Additional QC procedures may be necessary, depending on the purpose
      of the analysis performed with this method.

      10.10.1   Laboratory Duplicates — To determine precision
                associated with laboratory techniques, analyze two
                aliquots (Sect. 11.1.2) of a sample in which some
                analytes were detected in measurable quantities.
                Calculate the range (R) of concentrations measured for
                each duplicate pair:

                     R • Ci - Cg,

                where  CT represents the larger and,
                       C2 represents the smaller of the two
                       meas urements.

                Calculate percent relative range (RR) of duplicate
                analyses using the formula:

-------
                        RR
              R
              ^^^^

              C
100
         10.10.2
         10.10.3
where  R = range of concentrations measured, and
       C = mean concentration measured.
Generally, if RR is greater than 30*, precision is
inadequate, and laboratory techniques must be improved.

Field Duplicates — Analyze 10% of samples in which some
analytes were detected in measurable quantities to
indicate precision limitations imposed by sampling,
transport and storage techniques as well as laboratory
techniques.  If acceptable results are obtained from
analysis of field duplicates, analysis of laboratory
duplicates is usually not necessary.

Matrix Effects Determination — To indicate matrix
effects on method efficiency, accuracy and precision when
raw source waters or drinking water during treatment is
to be analyzed, analyze aliquots to which known amounts
of analytes have been added.  Because analytes may be
present in the unspiked aliquots, analysis of one or more
unspiked aliquots is necessary to determine initial
concentrations, which are then subtracted from concentra-
tions measured in spiked aliquots.  For each analyte the
amount added to determine matrix effects should exceed
twice the amount measured in unspiked aliquots.
11. PROCEDURE
    11.1 ANALYSIS PROCEDURES

         11.1.1 Initial conditions — Adjust the helium purge gas flow rate
                to 40 ± 3 mL/min.  Attach the sorbent trap to the purging
                device, and set the device to the purge mode.  Open the
                syringe valve located on the sample introduction needle of
                the purging chamber.

         11.1.2 Sample introduction and purging — Remove the plunger from a
                25-mL syringe and attach a closed syringe valve.  Open the
                sample or standard bottle, which has been allowed to come to
                ambient temperature, and pour the sample into the syringe
                barrel to just short of overflowing.  Replace the syringe
                plunger and compress the sample.  Open the syringe valve and
                vent any residual air while adjusting the sample volume to
                25.0 ml.  (Because this process of taking an aliquot impairs
                the integrity of the remaining sample, a second syringe
                should be filled at the same time, in case a second analysis
                is required.)  Add 2 to 20 uL of the spiking solution (Sect.
                7.5) of fluorobenzene in methanol through the syringe valve
                and close the valve.  Attach the syringe and its valve

-------
            assembly to the syringe valve on the purging device.   Open
            the syringe valves and slowly inject the  sample  into  the
            purging chamber.  Close both valves and purge the  sample  for
            11.0 ± 0.1 min at ambient temperature.  Because  temperature
            affects purging efficiencies of some analytes, purging
            chamber temperature must be controlled to maintain constant
            (approximately ± 2°C) temperature throughout calibration
            and sample analyses.  If laboratory temperature  is not
            controlled adequately, the purging chamber can be  placed  in
            a thermostatically controlled water bath.

     11.1.3 Desorption and data acquisition — At the conclusion  of
            purging, adjust the purge and trap apparatus to  the desorb
            mode, and initiate GC temperature programming, trap heating,
            and MS data acquisition.  Desorb for 4 min.  Transfer
            trapped sample components into the GC column by  heating the
            trap to 18QOC rapidly while it is backflushed with helium
            flowing at 20 to 60 mL/min.  (If the trap cannot be heated
            rapidly, use the GC column as a secondary trap by  cooling
            the column to < 30°C during desorption.)

     11.1.4 Sample chamber rinsing — During or after desorption  empty
            the purging chamber with the sample introduction syringe,
            and rinse the chamber with two 25-mL portions of reagent
            water.

     11.1.5 Trap reconditioning — After desorbing the sample  for 4 min,
            reset the purging device to the purge mode.  After 15 s,
            close the syringe valve on the purging device to begin gas
            flow through the trap.  After approximately 7 min, turn off
            the trap heater and open the syringe valve to stop gas flow
            through the trap.  When cool (<25°C), the trap is  ready,
            for the next sample.

     11.1.6 Termination of data acquisition — When sample components
            have eluted from the GC, terminate MS data acquisition and
            store data files on the data system storage device.   Use
            appropriate data output software to display full range mass
            spectra and appropriate extracted ion current profiles
            (EICPs).  If any ion abundance exceeds the system  working
            range, dilute the sample aliquot in the second syringe with
            reagent water and analyze the diluted aliquot.

11.2 IDENTIFICATION PROCEDURES CRITERIA — Tentatively identify a sample
     component by comparison of its mass spectrum (after background
     subtraction) to a reference spectrum in a collection.   Use the
     following criteria to confirm a tentative identification:

     11.2.1 The GC retention time of the sample component must be within
            £ s of the time observed for that same compound  when  a
            calibration solution was analyzed.  Calculate the  value of £
            with the equation:

-------
                        t - (RT)V3

                where RT = observed retention time (in seconds) of the
                compound when a calibration solution was analyzed.

         11.2.2 All ions that are present above 10% relative abundance in
                the mass spectrum of the standard must be present in the
                mass spectrum of the sample component and should agree
                within absolute 10%.  For example, if an ion has a relative
                abundance of 30% in the standard spectrum, its abundance in
                the sample spectrum should be in the range of 20 to 40%.

         11.2.3 Identification is hampered when sample components are not
                resolved chromatographically and produce mass spectra
                containing ions contributed by more than one analyte.
                Because purgeable organic compounds are relatively small
                molecules and produce comparatively simple mass spectra,
                this is not a significant problem for most method analytes.
                When GC peaks obviously represent more than one sample
                component (i.e., broadened peak with shoulder(s) or valley
                between two or more maxima), appropriate analyte spectra and
                background spectra can be selected by.examining EICPs of
                characteristic ions for tentatively identified components.
                When analytes coelute (i.e., only one GC peak is apparent),
                the identification criteria described in Section 11.2.2 can
                be met but each analyte spectrum will contain extraneous
                ions contributed by the coeluting compound.

         11.2.4 Structural isomers that produce very similar mass spectra
                can be explicitly identified only if they have sufficiently
                different GC retention times.  Acceptable resolution is
                achieved if the height of the valley between two isomer
                peaks is less than 25% of the sum of the two peak heights.
                Otherwise, structural isomers are identified as isomeric
                pairs.

12.  CALCULATIONS

    12.1 Complete chromatographic resolution is not necessary for accurate
         and precise measurements of analyte concentrations, if unique ions
         with adequate intensities are selected for EICPs.  For example,
         although two listed analytes, 1,1,2,2-tetrachloroethane and
         tetracnloroethene, were not resolved with the GC conditions used
         and produced mass spectra containing common ions, concentrations
         (Table 5) were calculated by measuring appropriate characteristic
         ions.

         12.1.1 With either the internal or external standard technique,
                calculate analyte concentrations with the equation:


                      Cx  .  Ax •  °s          .
                       x     AS .  RF  :v

-------
                where Cx =   analyte concentration  in  micrograms  per liter;
                      Ax =   integrated  ion abundance  of  a significant
                             characteristic ion of  the sample  analyte;
                      AS »   integrated "ion abundance  of  a significant
                             characteristic ion of  the standard  (either
                             internal or external),  in units consistent  with
                             those used for the analyte ion abundance;
                      RF   « response factor  (With  an  external standard,
                             RF » 1, because  the standard is the  same
                             compound as the  measured  analyte.);
                      Qs =   quantity of internal standard added  or  quantity
                             of external standard that produced As>  in
                             micrograms; and
                      V    » purged sample volume in liters.

         12.1.2 With the external standard technique,  As  is .a  point  on the
                concentration calibration curve and  is the same number as
                AX; QS is the quantity that produces As and also  is
                obtained from the concentration calibration curve.

         12.1.3 For each analyte, select a significant characteristic ion.
                When feasible, use the most intense  ion in the mass
                spectrum; when a less intense ion is more characteristic and
                sufficiently intense to provide necessary sensitivity, use
                that ion to avoid possible interferences.

13. PRECISION AND ACCURACY

    13.1 To obtain method efficiency data (Table 2)  and to indicate
         anticipated single laboratory accuracy and precision data (Table 5)
         for each listed analyte, seven 25-ml aliquots of each of two
         solutions of reagent water containing known amounts of analytes
         were analyzed with purge and trap procedures  and a packed column.
         One solution contained 16 ug of analyte per liter of  solution;  the
         other contained 1.6 ug/L.  Two direct Injections of appropriate
         volumes of secondary dilution standard were interspersed among
         purged aliquots.  To obtain the data in Table 5, one  aliquot of
         each of the two laboratory control standards  was randomly selected
         to be a solution with known true values of  analytes.  This  aliquot
         was treated as an external standard,  and the  other six aliquots of
         each of the two solutions were treated as  samples.

         13.1.1 Except for two listed analytes, mean method efficiency
                varied among analytes from 25.0% to  118.7%.  Those two
                analytes, l,2-dibromo-3-chloropropane  and bis(2-chloro-
                isopropyl) ether, are very inefficiently purged and  were not
                detected in aliquots containing 1.6  ug/L; mean method
                efficiencies for these two analytes when purged from 16  wg/L
                aliquots were 9.4% and 4.3%,  respectively (Table  2).
                Although for some applications these low efficiencies may
                result in unacceptably high detection  limits for  those

-------
       analytes, they can be measured with acceptable accuracy and
       precision when present at a concentration of 16 ug/L (Table
       5).

13.1.2 With these data, MDLs were calculated using the formula:


         HDL • Vl.l-  -0.99)     S-

where:
t(n-l, 1-  = 0.99) * Student's t value for the 99% confidence
                       level with n-1 degrees of freedom, where
                       n - number of replicates, and
                 s = standard deviation of replicate analyses.

-------
REFERENCES

1.  Glaser, 0. A., D. L. Foerst, G. D. McKee, S. A. Quave, and W. L. Budde,
    "Trace Analyses for Wastewaters," Environ. Sci. Techno!. 15, 1426, 1981.

2.  "Carcinogens - Working With Carcinogens," Department of Health,
    Education, and Welfare, Public Health Service, Center for Disease
    Control, National Institute for Occupational Safety and Health  ,
    Publication No. 77-206, Aug. 1977.

3.  "OSHA Safety and Health Standards, General Industry," (29CFR1910),
    Occupational Safety and Health Administration, OSHA 2206, (Revised,
    January 1976).

4.  "Safety in Academic Chemistry Laboratories," American Chemical  Society
    Publication, Committee on  Chemical Safety, 3rd Edition, 1979.

5.  "The Determination of Halogenated Chemicals in Water by the Purge and
    Trap Method," Method 502.1,  EPA 600/4-81-059,  U.S. Environmental
    Protection Agency, Office  of Research and Development, Environmental
    Monitoring and Support Laboratory, Cincinnati, OH, April 1981.

6.  "The Analysis of Aromatic  Chemicals  in  Water by the Purge and Trap
    Method,"  Method 503.1, EPA 600/4-81-057,   U.S. Environmental
    Protection Agency,. Office  of Research and Development, Environmental
    Monitoring and Support Laboratory, Cincinnati, OH, May 1980.

-------
Table 2.  Single Laboratory Method Efficiency  Data3 for  Purgeable
              Organic Compounds Measured with  GC/MS
Compound
chl oromethane

vinyl chloride

methyl ene chloride

1,1-dichloroethene

bromochl oromethane

trans-1 ,2-dichloroethene

chloroform

1,2-dichloroethane

1,1,1 -tri ch 1 oroethane

carbon tetrachloride

bromodi chl oromethane

1 , 1 ,2-tri ch loroethane

trichloroethene

benzene

chlorodibromomethane

cis-l,3-dich1oropropene

fluorobenzene
Measured
Ion
50

62

84

96

128

96

83

62

97

117

83

97

130

78

129

75

96
Rel.
Ret.
Timeb
0.10

0.14

0.22

0.28

0.30

0.31

0.44

0.50

0.59

0.62

0.66

0.79

0.84

0.88

0.88

0.90

1.00
True
Cone.
ug/L
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16 '
1.6
16
1.6
16
Mean Method
Efficiency

108.5

53.8
118.7
81.9
93.8
88.1
68.8
65.0
50. OC
98. 8C
93.8
92.5
62.5
55.6
93.8
90.6
106.2
110.0
81.3
80.6
.18.8C
50. QC
100.0
106.9
106.2
96.3
58.8
53.8
25. QC
70. Oc
100.0
95.6
Rel.
Std.
Dev., %

15.2

13.6
4.9
7.6
5.1
7.9
5.8
6.2
30. 5C
12. 6C
3.8
2.1
12.6
7.1
3.8
2.9
2.2
1.9
5.9
5.4
33. 5C
15.9C
3.0
1.2
2.9
2.2
6.0
11.3
31. 6C
11.8C
7.4
11.1

-------
                               Table 2.  (continued)
Compound
bromoform

1,1,2,2-tetrachloroethane

tetrachl oroethene

tol uene

chlorobenzene

1 ,2-dibromo-3-ch loropropane0"

4-bromof 1 uorobenzene

styrene

p-xylene

bis(2-ch1oroisopropyl) etherd

1,3-dichlorobenzene

1 ,2-di chl orobenzene

Measured
Ion
173

83

164

92

112

157

174

104

106

45

146

146

Rel.
Ret.
Timeb
1.10

1.29

1.31

1.42

1.52

1.70

1.82

1.93

1.97

2.08

2.19

2.20

True
Cone.
ug/L
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16

16
1.6
16
1.6
16
Mean Method
Efficiency
%
39.4
35.0
28.1
25.0
106.2
107.5
100.0
102.5
93.8
96.3
1-d
9.4C
93.8
93.8
68. 8C
86. 9C
100.0
103.1

4.3
75. OC
82. 5<=
81.3
79.4
Rel.
Std.
Dev %
7.0
6.3
5.9
6.3
0.0
4.6
3.2
1.5
3.5
2.2
lid
10.3C
5.4
4.0
29. 2C
13.3C
3.0
1.5

14.3
28. 3C
10.3C
5.9
4.2
a Except as noted, data were produced by purging  seven  aliquots of reagent
  water spiked with known amounts of  listed compounds;  calculations involved two
  direct injections.
b SC column: 1.8 m x 2 mm ID glass packed with  1% SP-1000 on 60/80 mesh
  carbopack B.  Program: 45°C for 4 min; 8°C/min  to 230°C.
  Retention time relative to fluorobenzene, which has retention time of  11.1 min
  under described GC conditions.
c Produced by analysis of six aliquots rather than seven.
d Compound is very inefficiently purged from water and  was not detected  in
  aliquots of 1.6 ug/L solution.

-------
Table  1.   Ion  Abundance Criteria for p_-Bromofluorobenzene




Mass                          Ion  Abundance  Criteria
50
75
95
96
173
174
175
176
177
15 to 40% of mass 95
30 to 60% of mass 95
Base Peak, 100% Relative Abundance
5 to 9% of mass 95
< 2% of mass 174
> 50% of mass 95
5 to 9% of mass 174
> 95% but < 101% of mass 174
5 to 9% of mass 176

-------
                              Table 3 (continued)
                                        Demonstrated  Acceptable
     Analyte                                 Storage  Time,
                                                 Days
2-chloroethylethyl ether                          27
2-chloroethylvinyl ether                          27
bis-2-chloroethyl ether                            9b
bis-2-chloroisopropyl  ether                       27


a These data were obtained by multiple analyses  of raw river water  and
  carbon-filtered chlorinated tap water to which known amounts  (0.20 to 0.50
  ug/L) of  listed analytes had been  added.  Some samples  were stored and
  analyzed  periodically over a 21-day period;  others,  over a 27-day period.
  Data from "The Determination, of Halogenated  Chemicals in Water by the
  Purge and Trap Method," Method 502.1, EPA 600/4-81-059,  U.S.  Environmental
  Protection Agency,  Office  of Research and Development,  Environmental
  Monitoring and Support  Laboratory, Cincinnati, OH, April 1981.

b Because of observed changes during storage,  this number is the maximum
  recommended  storage time.

-------
          Table 3.  Acceptable Storage Times3 for River and Drinking
                   Water Samples Containing Halogenated Aliphatic Analytes


                                       Demonstrated Acceptable
     Analyte                                Storage Time,
                                                Days

chloromethane                                    21
dichloromethane                                  27
carbon tetrachloride                             27
bromomethane                                      2b
dibromomethane                                   21
bromoform                                        27
bromochloromethane                               21
bromodichloromethane                             27
chlorodibromomethane                             27
dichlorodifluoromethane                          27
fluorotrichloromethane                           27

chloroethane                                     21
1,1-dichloroethane                               27
1,2-dichloroethane                               27
1,1,1-trichloroethane                            21
1,1,2-trichloroethane                            27
1,1,1,2-tetrachloroethane                        21
pentach1oroethane                                27
1,2-dibromoethane                                21

chloroethylene  (vinyl chloride)                   6b
1,1-dichloroethylene                             27
cis + trans-1,2-dichloroethylene                 27
cis-l,2-dichloroethylene                         21
1,1,2-trichloroethylene                          27
1,1,2,2-tetrachloroethylene                      27

1,2-dichloropropane                              21
1,3-dichloropropane                              21
1,2,3-trichloropropane                           21

3-chloroprop-l-ene  (ally!  chloride)               2b
1,1-dichloroprop-l-ene                           27
2,3-dichloroprop-l-ene                            °.
trans-l,3-dichloroprop-l-ene                      ]"
cis-l,3-dichloroprop-l-ene                       lu

1-chlorohexane                                    ]
chlorocyclohexane                                u
1-chlorocyclohex-l-ene                           Z7

-------
                             Table 4.  Storage Time Data3 for River and Drinking
                                  Water  Samples  Containing Aromatic Analytes
Analyte
  Percent Recovery From
  River Water on Day 1
Preserved*1      Unpreserved
Acceptable0 Storage Time
        In Days
  Drinking     Preserved
   Water      River Water
    Observed
    Average Decay
   Rate (%/Day) In
Preserved River Water
benzene
toluene
ethyl benzene
o-xylene
m-xylene
p-xylene
ethenyl benzene (styrene)
n-propylbenzene
isopropyl benzene
n -butyl benzene
sec-butylbenzene
ter -butyl benzene
1-methyl -4- 1 sopropylbenzene
1 , 2, 4 -tr1 methyl benzene
1 ,3,5-tr1methylbenzene
chlorobenzene
bromobenzene
o -d i ch 1 oro ben zene
m-di chl orobenzene
p-di chlorobenzene
1 ,2, 3-tri chlorobenzene
1,2,4-trlchlorobenzene
o-chlorotoluene
p-chlorotoluene
<*-trifluorotoluene
100
98
95
95
93
88
83
88
93
88
88
90
88
85
88
90
90
96
96
94
86
88
--
90
84
12
8
10
12
17
12
0
15
0
3
10
43
34
16
5
23
17
85
96
90
97
88
• •»
45
74
15
15
15
15
15
15,
0<«
15
15
156
15
15
6
15^
28
28
28
28
28
28
28
28
21
13
28
14
14
*••
__
—
_-
—
—
— •
•*••
—
~
M*»
• ••
	
26
26
26
26
26
26
26
—
26
26


1.4
1.6
1.6
2.3
4.2
2.2
1.5
3.2
2.2
10
2.3
3.0
2.1










a  Data  from  "The  Determination  of  Aromatic  Chemicals  1n  Water  by the Purge  and Trap  Method,"  Method 503.1,
   EPA 600/4-81-057,  U.S.  Environmental  Protection  Agency, Office of Research  and  Development,  Environmental
   Monitoring and  Support  Laboratory,  Cincinnati, OH,  May 1980,  and from  "The  Determination  of Halogenated
   Chemicals  in Water by the  Purge  and Trap  Method," Method  502.1, EPA  600/4-81-059,  U.S.  Environmental
   Protection Agency, Office  of  Research and Development, Environmental Monitoring and  Support Laboratory,
   Cincinnati, OH,  April 1981.
b  Preservation was accomplished by adjustment  of sample  pH  to  2.
c  Data  were  obtained by multiple analyses of carbon-filtered tap water and  preserved river  water to which
   known amounts (0.40 or  0.50 pg/L) of  listed  analytes had  been added.    Mean recovery of analyte was ^ 80%
   except as  noted.
d  Styrene, which  reacts with free  chlorine  was not detected in chlorinated  drinking  water.
f  Mean recovery of 78$.
f Mean recovery of 75%.

-------
Table 5.  Anticipated Accuracy and Precision Data with Method
            (Single Laboratory and Single Operator)
Compound
chloromethane

vinyl chloride

methyl ene chloride

1 , 1-di chl oroethene

bromochl oromethane

trans-l,2-dichloroethene

chloroform

1,2-dichloroethane

1,1, 1-tri chl oroethane

carbon tetrachloride

bromodi chl oromethane

1 , 1 ,2-tri chloroethane

tri chl oroethene

benzene

chl orodibromome thane

cis-l,3-dichloropropene

fluoro benzene
Measured
Ion
50

62

84

96

128

96

83

62

97

117

83

97

130

78

129

75

96
Rel.
Ret.
T1meb
0.10

0.14

0.22

0.28

0.30

0.31

0.44

0.50

0.59

0.62

0.66

0.79

0.84

0.88

0.88

0.90

1.00
True
Cone.
yg/L
Oe
16
Oe
16
1.6
16
1.6
16
1.6
16
1.6,
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16

Mean
Observed
Cone., yg/L

16.5

18.2
1.6
15.8
1.6
16.3
1.6
15.6
1.3
15.4
1.6
16.0
1.6
16.0
1.6
16.2
1.6
16.0
1.6
15.8
1.3
15.9
1.6
16.0
1.6
16.1
1.6
15.0
1.3
15.0

Std.
Dev.
yg/L

2.7

2.5
0.08
0.3
0.08
1.8
0.08
1.4
0.5
2.2
0.06
0.32
0.1
1.7
0.04
0.8
0.04
0.4
0.09
0.5
0.3
2.5
0.05
0.2
0.06
0.5
0.1
2.0
0.5
1.8

Rel.
Std.
Dev.,%

16.6

13.9
4.8
2.1
4.9
10.8
5.2
9.2
37.9
14.5
3.8
2.0
6.7
10.5
2.4
4.9
2.3
2.4
5.5
3.2
23.1
15.9
3.3
1.1
4.0
3.1
6.5
13.6
36.9
11.7

Mean Method
Accuracy

103.1

113.8
100
98.8
100
101.9
100
97.5
81.3
96.3
100
100
100
100
100
101.3
100
100
100
98.8
81.3
99.4
100
100
100
100.6
100
93.8
81.3
93.8

Method
Detection
Limit0, yg/L

9.2

8.5
0.25
—
0.27
—
0.28
—
1.7
—
0.20
-_
0.35
—
0.13
__
0.13
_«.
0.29
__
1.0
....
0.18
«._
0.21
— «.
0.34
mmmm
1.8
— .*


-------
                                                 Table 5.   (continued)
Compound
bromoform

1 ,1 ,2,2-tetrachloroethane

tetrachloroethene

toluene

chlorobenzene

1 ,2-d1bromo-3-chloropropane('

4-bromofluorobenzene

styrene

p-xylene

b1s(2-chloro1sopropyl) ether**

1 ,3-dichlorobenzene

1,2-di chlorobenzene

Measured
Ion
173

83

164

92

112

157
•
174

104

106

45

146

146

Rel.
Ret.
T1meb
1.10

1.29

1.31

1.42

1.52

1.70

1.82

1.93

1.97

2.08

2.19

2.20

True
Cone.
V9/L
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
1.6
16
Mean
Observed
Cone., ug/J.
1.5
15.5
1.6
15.5
1.6
15.8
1.6
15.9
1.6
15.7
__d
15.7
1.5
15.6
1.2
15.8
1.6
16.1
_-d
15.1
1.2
15.2
1.5
15.7
Std.
Dev.
uq/L
0.1
2.0
0.08
1.8
0.02
0.4
0.02
0.3
0.03
0.6
_.d
3.0
0.09
1.1
0.3
1.6
0.05
0.2
_.d
2.3
0.3
1.5
0.09
1.3
Rel.
Std.
Dev.,%
6.5
12.7
5.3
11.5
1.3
2.4
1.4
1.6
1.6
3.6
__d
19.1
5.6
7.3
29.0
10.4
3.4
1.0
__d
15.1
28.0
9,8
6.2
8.3
Mean Method
Accuracy
%
93.8
96.9
100
96.9
100
98.8
100
99.4
100
98.1
__d
98.1
93.8
97.5
75.0
98.8
100
100.6
-_d
94.4
75.0
95.0
93.8
98.1
Method
Detection
L1m1tc,ug/L
0.34
• M»
0.28
__
0.07
__
0.08
__
0.09
— —
-d
«
10. ld
0.29
--
1.3
—
0.18
_—
_.d
j
8.6d
1.3
-
0.30
-
a Produced by analysis of seven allquots of reagent  water spiked with known amounts of listed compounds;
  calculations based on external standard technique.   Two allquots were treated as standards; five
  allquots were treated as samples.
b GC column:  1.8 m x 2 mm ID glass packed with  1% SP-1000 on 60/80 mesh Carbopack B,  Retention times
  relative to fluorobenzene.
c Minimum concentration that can be measured with  995t confidence that reported value Is greater than
  zero.1
d Compound is inefficiently purged and was not detected 1n 1.6 ug/L solution; MDL was calculated from
  analysis of 16 yg/L solution.
e Compound not analyzed at concentration of 1.6  ug/L.

-------