United States
Environments! Protection
Agency
Environmental Sciences Research
Laboratory
Reeearch Triangle Park NC 27711
EPA-600/2-79-121
July 1979
Research and Development
Development of an
SO2 Monitor for
Mobile Sources

-------
                RESEARCH REPORTING SERIES

Research reports of the Office of Research and Development, U S. Environmental
Protection Agency, have been grouped into nine series. These nine broad cate-
gories were established to facilitate further development and application of en-
vironmental technology  Elimination of traditional grouping was consciously
planned to foster technology transfer and a maximum interface in related fields.
The nine series are:

      1.   Environmental  Health Effects Research
      2.   Environmental  Protection Technology
      3.   Ecological Research
      4.   Environmental  Monitoring
      5.   Socioeconomic Environmental Studies
      6.   Scientific and Technical Assessment Reports (STAR)
      7.   Interagency  Energy-Environment Research and Development
      8.   "Special" Reports
      9.   Miscellaneous Reports

This report has been assigned to the ENVIRONMENTAL PROTECTION TECH-
NOLOGY series. This series describes research performed to develop and dem-
onstrate instrumentation, equipment, and methodology to repair or prevent en-
vironmental degradation from point and non-point sources of pollution. This work
provides the new or improved technology required for the control and treatment
of pollution sources to meet environmental quality standards.
This document is available to the public through the National Technical Informa-
tion Service, Springfield, Virginia 22161.

-------
                                                 EPA-600/2-79-121
                                                 July 1979
           DEVELOPMENT OF AN SO. MONITOR
                FOR MOBILE SOURCES
                         by
     Darrell E. Burch, Pamela S. Marrs Davila,
       Francis J. Gates and John D. Pembrook
   Ford Aerospace and Communications Corporation
               Aeronutronic Division
          Newport Beach, California  92663
               Contract No. 68-02-2448
                   Project Officer

                   Roy Zweidinger
Emissions Measurement and Characterization Division
    Environmental Sciences Research Laboratory
  Research Triangle Park, North Carolina  27711
    ENVIRONMENTAL SCIENCES RESEARCH LABORATORY
        OFFICE OF RESEARCH AND DEVELOPMENT
     U.  S.  ENVIRONMENTAL PROTECTION AGENCY
   RESEARCH TRIANGLE PARK, NORTH CAROLINA  27711

-------
                                DISCLAIMER
This report has been reviewed by the Environmental Sciences Research Laboratory,
U.S. Environmental Protection Agency, and approved for publication.  Approval
does not signify that the contents necessarily reflect the view and policies
of the U.S. Environmental Protection Agency, nor does mention of trade names or
commercial products constitute endorsement or recommendation for use.

-------
                                 ABSTRACT

     An instrument has been designed and built to monitor the concentration
of S02 in the exhaust of mobile sources.  A grating assembly disperses the
ultraviolet energy from a deuterium-arc source and passes five narrow
spectral intervals.  Three of the intervals (set A) coincide with strong
absorption features in the S02 band near 3000 A; the other two intervals
(set B) coincide with weak absorption.  A spinning reticle alternately
transmits energy passing through set A and B to a photomultiplier detector.
A dc output signal proportional to the concentration of S(>2 in the sample
cell is produced by appropriate electronics.

     A pump and manifold assembly permit operation over a wide range of
sample flow rates.  At high flow rates, the 90% sample turn-over time is
approximately 0.7 sec.   The sample cell contains a multiple-pass optical
system adjusted to 12 passes, giving a sample path length greater than
5 m.  The rms noise level corresponds to approximately 0.05 ppm of S02
when the electronic time constant is 1 sec.

     This report was submitted in fulfillment of Contract No. 68-02-2448
by Ford Aerospace and Communications Corporation, Aeronutronic Division
under the sponsorship of the U.S. Environmental Protection Agency.  This
report covers the period September 30, 1976 to January 31, 1978, and work
was completed as of September 30, 1978.
                                     1X1

-------
                                  CONTENTS

     \
Abstract                                                                iii

Figures                                                                  v

Abbreviations and Symbols                                               vi




1.   Introduction                                                        1

2.   Conclusions                                                         3

3.   Recommendations                                                     5

4.   Layout and Optical Diagram                                          7

5.   Sample Cell                                                        15

6.   Pumps and Manifold                                                 22

7.   Spectroscopic Principles of Detection                              26

8.   Electronic Processing of Detector Signal                           31

9-   Performance, Calibration, and Discrimination                       33

-------
                                  FIGURES

Number                                                                   Page

1       Photograph of the Optics Assembly, the Manifold
        Assembly, the Pump Assembly, and the UV Source
        Power Supply                                                       8

2   '    Photograph of the Optics Assembly with Front Hinged
        Panels Closed                                                      9

3       Photograph of the Optics Assembly with Front Hinged
        Panels Opened                                                     10

4       Optical Diagram of the S02 Monitor                                11

5       Sectional View of the Sample Cell                                 17

6       Multiple-Pass Optical System                                      22

7       Flow Diagram of Pump Assembly                                     24

8       Flow Diagram of Manifold Assembly                                 25

9       Spectrum of S02 from 2600 to 3300 Angstroms                       27

10      Partial Spectrum of S02 Band Showing Spectral Slits
        A and B                                                           28

11      Block Diagram of Signal-Processing Electronics                    29
                                     VI

-------
                         ABBREVIATIONS AND SYMBOLS
AGC      —automatic gain control

cfh      —cubic feet per hour

cfm      —cubic feet per minute

cm       —centimeter

1pm      —liters per minute

m        —meter

ppm      —parts per million by volume

T        —transmittance at a particular wavennumber (or wavelength) that
           would be measured with a spectrometer having infinite resolving
           power

T.       —average transmittance of the absorbing gas (S0_) in spectral
           intervals designated by A

T        —average transmittance of the absorbing gas (S0_) in spectral
           intervals designated by B

V_       —voltage component of amplified detector signal at carrier frequency,
           360 Hz

V        —voltage component of amplified detector signal at alternator
           frequency, 30 Hz

V'       —Va/Vc, normalized voltage.  Normalization is made so that V' =
           1 when there is 100% modulation of the beam at 30 Hz
                                      Vll

-------
                                 • SECTION 1

                                INTRODUCTION
     Most of the sulfur in gasoline is oxidized upon combustion in an engine
to S02<  Concentrations of this gas in raw exhaust are typically between 5 ppm
and 50 ppm.  At these concentrations, the S02 is not considered a serious
pollutant unless it is oxidized further to form 803, which reacts rapidly
with.-H2-0- to form toxic sulfuric acid.  Catalytic converters are known to
oxidize some of the S0£ to the more harmful 803.  Unfortunately, the process
by which the 802 is converted to 803, and ultimately to acid, is difficult
to study quantitatively because there is no convenient, real-time method of
measuring the acid concentration in the exhaust.  It has been found, however,
that much can be learned about the behavior of the catalytic converters and
about the entire 802 conversion process by monitoring the concentration of
802 in the exhaust.

     The instrument described herein has been designed and built for this
purpose, to monitor in real-time the 802 concentration in engine exhaust.
The low concentration requires that the instrument be quite sensitive and
that the discrimination against other exhaust gases be good.  Rapid changes
in the 802 concentration have been observed previously, and it is important
that these changes be recorded.  This leads to another important requirement,
quick response, for the monitor.

     The instrument achieves high sensitivity by making use of the very
strong 802 absorption band near 3000 A while employing a multiple-pass optical
system in the sample cell to obtain a long absorption path in a small volume.
An instrument built previously by usl for the EPA to monitor 802 ^n stac^
exhaust made use of the infrared absorption band of 802 near 4 ym.  This
infrared band is too weak to provide the high sensitivity required for the
present instrument.  Excellent discrimination has been achieved by employing
a correlation method that makes use of the structure in the uv absorption
band.  A low-volume sample cell designed for high gas flow rates used in
conjunction with high-capacity pumps makes the quick response possible.
     1.   Burch,  D.  E., and D. A.  Gryvnak.   "Infrared Gas Filter Correlation
         Instrument for In-Situ Measurement of Gaseous Pollutants."  Prepared
         by Philco-Ford Corporation for EPA under Contract No.  68-02-0575.
         EPA Report No.  EPA-650/2-74-094.   Also,  Burch,  D. E.,  and D.  A.
         Gryvnak.   "Cross-Stack Measurement of Pollutant Concentrations
         Using Gas-Cell Correlation Spectroscopy."  Chapter 10 of Analytical
         Methods Applied to Air Pollution Measurements,  Stevens, R.  K. and
         W.  F.  Herget, (eds.).   Ann Arbor,  Ann Arbor Science Publishers Inc.,
         1974.

-------
The important design goals are summarized as follows:

   Sensitivity ranges, full scale:  5, 10, 20, and
      100 ppm.

        Higher concentrations can be measured by
        decreasing the number of passes of the
        multiple-pass cell.

   Noise-equivalent-concentration:  =0.05 ppm

   Sample turn-over time (90%):  =0.5 sec at maximum flow rate.

   Interference by other gases in automotive exhaust should not
      produce significant error in the S0« measurement.

-------
                                 SECTION 2

                                CONCLUSIONS
     It has been demonstrated that an instrument that makes use of the strong
S02 absorption features near 3000 A can be designed and built with the sensi-
tivity and accuracy required to measure the concentration of S02 in automotive
exhaust.  Good discrimination against interference by other gases in the
exhaust can be achieved by employing the correlation technique used in the
instrument described herein.  This technique, when used properly, provides
very little sensitivity to changes in the reflectivity of mirrors, the trans-
mittance of windows, or variations in source intensity.  Electronic circuits
developed and used previously by us to process the detector signals of gas-
filter correlation instruments work equally well with the present instrument,
which employs a grating polychromator with a spinning reticle in place of a
gas-filter cell.

     The minimum detectable concentration is limited by photoelectron noise
generated in the photomultiplier.  The noise level measured at the output of
the photomultiplier is approximately proportional to the square-root of the
radiant energy in the ultraviolet beam incident on the photomultiplier.  It
follows that increasing the energy level by such means as increasing the
source intensity or improving mirror reflectivity would result in an increase
in S/N, the signal-to-noise ratio, that is proportional to the square-root
of the energy level.

     The maximum S/N for a given S02 concentration is obtained in the present
instrument when the multiple-pass sample cell is adjusted to 12 passes.  The
necessity for this low number of passes is due to the low effective reflec-
tivity (between 0.80 and 0.85) of the uv-enhanced aluminum coating on the
mirrors of the multiple-pass optical system.  The actual reflectivity may be
higher than this value, but each reflection adds slightly to the distortion
of the images.  This additional distortion has the effect of lowering the
reflectivity by decreasing the amount of energy entering the entrance slit of
the grating assembly.  Mirrors with a dielectric reflective coating can be
made with better reflectivity than the mirrors used near 3000 A, the spectral
region of interest.  The reflective coating that is more efficient in the uv
was not used because its very low reflectivity in the visible makes it diffi-
cult to align the optics with visible light.

     Short turn-over times for samples in the sample cell can be achieved by
flushing the gas in a low-turbulent stream perpendicular to the monitor beam.
If the sample cell is shaped properly and the mirrors and windows are mounted
properly, the noise produced on the detector signal by the turbulence is not

-------
significant.  Little or no error in sampling is produced by adsorption of SC>2
on the sample cell surfaces because they are coated with non-reactive Kel-F.
Teflon coatings adsorbed S02 more readily than Kel-F.  Anodized aluminum sur-
faces should not be exposed to S02 samples because of the very rapid adsorption.

-------
                                 SECTION 3

                              RECOMMENDATIONS
     The instrument described herein has been designed and built to meet a
combination of several desirable specifications as a monitor for S02 in mobile
sources.  Some of the important features are:  relatively small size, short
response time, variable sample flow rate, very good discrimination, good
accuracy, low minimum detectable concentration, and wide dynamic range.
Specifications for some of these features necessarily have been sacrificed
to improve the specifications of other features.  Thus, any S0£ monitor to
be patterned after this one undoubtedly should be modified according to the
specific types of measurements to be made with it.  A few design changes have
become apparent as a result of assembling and testing the present instrument,
but even more can be learned by using it for a few weeks under normal testing
conditions.

     Without changing the basic design, a few relatively minor changes could
be made to improve the signal-to-noise ratio by increasing the amount of
ultraviolet energy reaching the detector.  The biggest gains can be made by
employing mirrors with a dielectric reflective coating that has a much higher
reflectivity than the uv-enhanced aluminum presently being used.  This change,
however, would complicate the visual optical alignment because the dielectric
mirrors reflect very poorly in the visible.  This disadvantage could be over-
come by employing mirror mounts that make it possible to remove a mirror and
replace it with one of the same radius of curvature without losing optical
alignment.  Mirror mounts could be adjusted with mirrors having good visible
reflectivity; the mirrors then could be replaced with ones having good uv
reflectivity.  Some of the flat mirrors should be replaced by quartz prisms
that are oriented to produce total internal reflection, which can be quite
efficient.  The prisms should be anti-reflection coated on the surfaces
through which the beam passes.

     By enlarging the cross-sectional area of the sample cell, the multiple-
pass mirrors can be enlarged, thus allowing more energy to be transmitted.
The sample cell could be lengthened to increase the electrical signal produced
by a given S02 concentration.  Either of these two changes necessarily in-
creases the sampling volume, which increases either the necessary flow rate
or the sample turn-over time, or both.  A grating assembly with longer focal
length mirrors and/or a larger grating might be required to take full advan-
tage of larger mirrors in the sample cell.

     Some gain in signal-to-noise ratio and instability can probably be
achieved by replacing the side-looking photomultiplier with an end-looking

-------
one that has a higher quantum efficiency.  Stability could also probably be
improved by employing a lens between the exit grid and the photomultiplier so
that the grating is focused on the photocathode.  This would reduce some of
the effect of the inherent non-uniform sensitivity of side-looking photo-
cathodes.

     The size and complexity of the instrument can be decreased if the speci-
fications are relaxed on either the minimum detectable concentration, the
discrimination or the sample turn-over time.  As an example, the grating
assembly with the grid that produces relatively high spectral resolution
could be replaced with a pair of uv interference filters.  One filter would
transmit a spectral region that includes most of the strong absorption by
the entire S(>2 band; the other would transmit an adjacent spectral interval
in which the S02 absorption is weak.  By alternating the filters in the
monitoring beam and comparing the transmitted energy levels, the S02 con-
centration could be determined.  This simplified method makes use of the
absorption by most of the absorption band, but it does not make use of the
structure in the band (see Figures 9 and 10), with the result that the
discrimination is reduced and the zero-setting of the instrument is more
subject  to drifts caused by variations in source intensity or by changes in
the temperatures of the filters or other optical components.

     Gas-filter cells used in conjunction with an interference filter should
also be  considered as a simpler method.  However, gas-filter correlation
techniques are not expected to provide discrimination and stability as good
as that  provided by the grating assembly in the present instrument.

     The flow through the sample cell is not as free of mixing as was expected
and should be investigated further to determine ways of decreasing the turn-
over time.  Different shapes for the inside of the sample cell should be
tested at typical sample flow rates.  The gas flow should be kept parallel to
the faces of the multiple-pass mirrors (i.e. perpendicular to the monitoring
beam) if high flow rates are required.  This avoids excessive turbulence that
may be produced by gas flowing around the mirrors.

-------
                                 SECTION 4

                         LAYOUT AND OPTICAL DIAGRAM
     Figure 1 illustrates the pump assembly, the manifold assembly, and the
optics assembly.  The pump assembly is connected to the manifold assembly
with stainless steel tubing, and the manifold is connected to the inlet and
outlet of the aluminum sample cell with the  same  type  of  tubing.  The  uv source
power supply is also shown in Figure 1.  Compressed gas enters the manifold
after particulates are removed by a filter mounted on the pump assembly.  A
wide range of sample gas flow rates is obtainable with the two pumps mounted
on the pump assembly and the flow-meters and controls available on the mani-
fold assembly.  The gas pressure is indicated at key points in the flow path
by pressure gauges.  After leaving the manifold, the gas flows into the sample
cell and through the sampling volume where the monitoring beam is perpendic-
ular to the direction of the gas flow.  The gas absorbs ultraviolet radiant
energy from the monitoring beam and then exits from the sample cell.  The
sample gas is then exhausted through the manifold or passed on to other
instruments for further analysis.

     Figure 2 is a front view of the optics assembly with the front hinged
panels closed and with the available components and controls labeled.
Figure 3 is a front view of the optics assembly with the front hinged panels
opened and with components and controls labeled appropriately.  These figures
are discussed in more detail in this section.

     Figure 4 is an optical diagram of the S02 monitor.  Ultraviolet energy
from the deuterium arc source is focused by lens LI onto the field lens L2,
which serves as the entrance window of the sample cell.  The optical diagram
within the sample cell is too complex to illustrate completely for 12 passes
of the cell.  The distance between mirror C2 and either mirror Cl or C3 is
42.7 cm, giving a 512 cm path for 12 passes.  The number of passes can be
changed by integral multiples of 4 by use of the adjustment screws located
behind mirrors Cl and C3.  Mirrors Cl, C2, and C3 are bonded to stainless
steel back-up plates with the appropriate holes and grooves required for the
adjustments.

     After passing through the multiple-pass optical system in the sample
cell, the beam emerges through the exit window, lens L3.  The beam is then
incident on flat mirror XI, which is mounted 45° from the vertical.  Mirror XI
rotates the beam 90° and directs it downward to flat mirror X2.  Mirror X2
is mounted 45° from the vertical, and this mirror rotates the beam 90° and
directs it toward spherical mirror X3.  Before entering the sample cell, the
central ray of the beam is 15.0 cm above the baseplate of the optics assembly.

-------

Figure 1.  Photograph of the optics assembly, the manifold assembly,
           the pump assembly, and the UV source power supply.

-------
                                                     TIME CONSTANT SWITCH
c
     ANALOG
     PANEL
     METER

     DIGITAL
     PANEL
     METER
                         DIGITAL OUTPUT
                         JACK
FIXED ANALOG
OUTPUT CONNECTOR
UV SOURCE POWER
CONNECTOR

      UV SOURCE
      POWER SWITCH
                                   VARIABLE ANALOG
                                   OUTPUT CONNECTOR
                                          ZERO  BALANCE  SWITCH
                                          AND POTENTIOMETER
                                                SAMPLE  CELL
                                                PRESSURE  GAUGE
LINE CORD
CONNECTOR

              Figure 2.  Photograph of the optics assembly with front hinged jpaneIs closed.

-------
       CALIBRATION CELL
       LIFTING ROD
INDIVIDUAL
RANGE POTENTIOMETERS

INDIVIDUAL
DIGITAL METER
SPAN POTENTIOMETERS
                                                          GRATING
                                                          ASSEMBLY
                  TEMPERATURE
                  CONTROLLER
           ± 15 VDC
           POWER SUPPLY
                                                                                               MOTOR TRANSFORMER

                                                                                               HIGH VOLTAGE
                                                                                               PM POWER SUPPLY
                                        HV ADJ
                                        POTENTIOMETER
                                       REFERENCE CARD
                                                                                                METER FUNCTION
                                                                                                SWITCH
                                                                                                   MASTER SPAN
                                                                                                   POTENTIOMETER
                         RANGE CARD
MAIN ELECTRONICS CARD
SIGNAL
INPUT
POTENTIOMETER
                       Figure 3.   Photograph  of  the  optics  assembly with  front hinged
                                  panels  opened.

-------
             r
                                                                                    Cl   (")
ARC
                                                 SAMi'I I. Cl 1.1.		-
                                                                                                        SAMJM.I  on n;orrn-:)
J
1-


-- PI i r %
	 	 	 ~~ ~~ \
\==^:~~~~^
•UAT r v, ~- 	 Q
f 	 * "H
•-K),0'.
1 30 r:x-









                                                1KAT1NC ASSEMBLE-
                                                                                       GRID
           1— SPINNING
:.--+_ T    RETICLE
                                  Figure 4. Optical diagram of the SC>2 monitor.

-------
Upon emerging from the sample cell,  the central  ray  is  11.9  cm above the base-
plate.  Mirrors XI and X2 lower the  beam 5.1 cm;  thus the beam is  6.8 cm above
the baseplate of the optics assembly after reflection from mirror  X2.   Mirror
X3 forms an image of the source on the 0.3 mm wide by 2.3 mm high  entrance
slit of the grating assembly.

     Spherical mirror Gl collimates  the beam of  energy  after it has  passed
through the entrance slit.  The combination of concave  mirror G2 (f  = 61 cm)
and convex mirror G3 (f = -53.2 cm)  focus the radiant energy dispersed by the
grating onto the grid.  The dispersion at the focal  plane where the  grid is
located is approximately three times as great as  it  would be if mirrors G2
and G3 were replaced by a single concave mirror  identical to mirror  Gl
(f = 40.6 cm).  This increased dispersion makes  it possible  to use a grid
with wider spacings between the openings.  In the optical diagram  of Figure 4,
energy of only a single wavelength is represented in the path from the
grating to mirror G2 to the grid.  Energy of a wavelength slightly different
from the wavelength represented is diffracted from the  grating at  a  slightly
different angle and is therefore brought to a focus  beside the point of focus
of the represented wavelength.  The  grid is a thin piece of  stainless  steel
shim stock with five slits cut in it.  Two 1.5 mm wide  by 4.6 mm high  slits
pass the two B spectral intervals in the S02 absorption band where weak
absorption occurs.  The three 1 mm wide by 4.6 mm high  slits pass  the  three A
spectral intervals in the S02 absorption band where  relatively strong
absorption occurs.  After passing through the slits, the radiant energy is
chopped at 360 Hz>and energy from either the three A intervals or  the  two B
intervals is incident on the CsSb cathode of the  photomultiplier,  which has a
modified S-5 spectral response.  The side-looking photomultiplier  is sup-
ported by a mount attached to the grating assembly and  extends into  the
cylindrical chopper.  The photomultiplier is centered inside the cylinder, and
the 24 mm wide x 8 mm high photocathode is perpendicular to  the beam of
radiant energy.  A reticle that functions both as a  chopper  and as a 30 Hz
alternator is bonded to the cylindrical chopper.  The reticle chops  the incom-
ing signal at 360 Hz and also alternately passes  the three A intervals,  then
the two B intervals at 30 Hz.  The reticle slits  are aligned with  the  grid
slits and are oversized so that this spinning component will not contribute
significantly to instabilities, noise, or zero drifts.

     The photomultiplier dynode chain has a ten-stage linear configuration.
The signal from the photomultiplier  is processed  electronically to produce
a dc output signal proportional to the concentration of S02  sample gas.   The
signal-processing electronics are described in detail in Section 8.
     A calibration cell that is filled with a mixture of  S02 and N2  can be
moved easily into the beam between mirror Nl and the entrance window to the
sample cell.  The calibration cell is not shown in Figure 4.  The  optical
pathlength through the calibration cell is approximately  4 mm.  The  cell can
be filled to any desired pressure up to 1 atmosphere.  Valves and  fittings
fastened permanently to the calibration cell make it convenient to change
the amount of gas in it without removing it from the instrument.

     Also not shown in Figure 4, but included with the instrument, is a
permanently mounted tungsten-iodide bulb.  The tungsten-iodide bulb  is mounted
                                     12

-------
below the beam between lens LI and mirror Nl  and provides visible  light  to
align the optical components in  the  sample  cell and  in  the  grating assembly.
The deuterium arc source emits very  little  visible energy,  making  it  difficult
to align optics, particularly after  the beam  has been reflected  a  few times.
A flat mirror, N2, and a lens, L5, are placed directly  above  the bulb and
in the path of the ultraviolet energy between lens LI and mirror Nl.   Thus,
from the movable mirror, the beam of visible  energy  traverses exactly the
same path as the ultraviolet energy  from the  arc source.

     A small window on the end of the sample  cell opposite  the entrance  and
exit windows makes it possible to view images formed on mirror C2  of  the
multiple-pass system to determine the number  of passes  of the monitoring beam.
A prism is mounted over the viewing  window  for viewing  the  images  on  mirror
C2 from the rear of the optics assembly.  The number of passes can be adjusted
from outside the sample cell without opening  the cell or disturbing the
sample.

     Figure 2 is a front view of the optics assembly.   A digital and  an  analog
panel meter are mounted on the large, front hinged panel.   The analog panel
meter displays either the dc output  signal, the input signal,  the  reference
signal, or the carrier level, depending on  the position of  the meter  function
switch mounted on the main electronics card.   The dc output signal is always
displayed on the digital panel meter regardless of the  position  of the meter
function switch.  The analog dc  output signal is monitored  by connecting a
strip-chart recorder to either the fixed or the variable output  connector
mounted on the lower panel.  The output of  the fixed output connector is
10 volts for a full-scale reading.   The full-scale output of  the variable
output connector can be adjusted from about 1 millivolt to  10  volts by
adjusting a potentiometer and the recorder  gain switch  on the  main electronics
card.

     The range switch mounted on the front  hinged panel is  used  to select
either the 5, 10, 20, 50, or 100 ppm range.  When the range switch is in
the off position, the dc output  of the instrument is zero.  The  time  constant
switch is also mounted on the front  hinged  panel; time  constants of 0.1, 0.3,
1.0, 2.0, and 5.0 seconds are available.  A 10-turn  zero-balance potentiometer
and a zero switch are used in combination to  electronically cancel any
residual 30 Hz modulation on the carrier when S02 is not present in the  sample
gas.

     AC power is supplied to the electrical components  by connecting  a line
cord to the "AC IN" male connector and by switching  the power  switch  to  the
on position.  When the power switch  is in either the standby  or  the on
position, power is supplied to the heaters  bonded to the sample  cell  and to
the power resistors mounted on the sample cell inlet and outlet  fittings.
After the cell has stabilized at 55°C, the  cell can  be maintained  at
this temperature with the power  switch in the standby position while  other
electrical components are inoperative.  This  feature allows the  cell  to  be
maintained at 55°C when tests are not being performed and thus the operator
does not have to wait 2 to 3 hours for the  cell temperature to stabilize each
time before data are taken.  The  source power  switch  located on the lower front
panel is used to turn the UV source  fan on  after the UV source power  supply
                                      13

-------
is connected to the source power outlet mounted on the lower front panel.
This outlet was installed as a safety feature to prevent the source from being
turned on when the cooling fan is not operating.

     Figure 3 is a front view of the optics assembly with the front hinged
panels open.  The main electronics card and the range card are mounted on
the inside surface of the front hinged panel.  The reference card is mounted
on the inside surface of the right side panel.  The high-voltage power supply
for the photomultiplier, the chopper motor transformer, the +15 Vdc power
supply, and the temperature controller are mounted on the power supply chassis.
The chopper motor is mounted on the stainless steel grating assembly and is
located below the power supply chassis.  The bottom of the sample cell is
secured to the baseplate of the optics assembly with two screws.  The cell is
also secured to the partition that separates the grating assembly from the
sample cell.  Figure 4 is a top view of the optics assembly showing the loca-
tions of the grating assembly, the sample cell, the chopper motor, and the
photomultiplier.

     Both the sample cell and the grating assembly rest on spacers that hold
these two components approximately 1.5 cm above the base plate.  The spacers
under the sample cell provide insulation to minimize the flow of heat from
the sample cell to the base plate.  The three spacers under the grating assem-
bly are arranged in such a way as to not put any stress on the grating assem-
bly if the main base plate is distorted slightly.
                                     14

-------
                                 SECTION 5

                                SAMPLE CELL
REQUIREMENTS AND GENERAL FEATURES

     The two specifications for the instrument that are important factors in
the design of the sample cell are the high sensitivity and the quick turn-
over of gas within the sampling volume.  The gas flow through the sampling
section may be varied, depending on the type of study being performed.  If the
concentration of SC>2 in the gas entering the gas line at maximum flow rate
changes suddenly from one concentration level A to a second concentration
level B, it is desired that the concentration within the sampling volume will
have changed 90% of the way from level A to level B within 0.5 seconds after
the concentration starts to change within the sampling volume.  This excep-
tionally good time resolution requires a gas flow system with a minimum
amount of mixing.  The flow rate must be sufficiently high to displace all of
the gas in the sampling, volume quickly enough to allow for a certain amount
of mixing that can not be avoided.  Turbulence produced by the high flow rate
must not produce significant noise on the signal from the photomultiplier
used to detect the monitoring beam.

     The second difficult specification calls for the signal-to-noise ratio,
S/N to be> 2 for an S02 concentration of 0.1 ppm.  This high sensitivity
requires that the monitoring beam pass through at least a few meters of
sample gas.  Thus, this imposes certain minimum limits on the size of the
sampling volume.  This limitation, in turn, requires a relatively high flow
rate of the sample gas in order to displace the gas within the sample volume
in the required time.  The long path of the monitoring beam is obtained by
employing a multiple-pass optical system in the sample cell.  The sampling
volume is made as small as practical while maintaining the mirrors in the
multiple-pass optical system large enough to pass an adequate amount of
ultraviolet energy through the monitoring system.
                                     15

-------
CELL CONSTRUCTION AND GAS FLOW

     A cross-sectional view of the sample cell is shown  in Figure  5.   The
sample cell is 42.7 cm long in the dimension perpendicular to  the  plane  of
the section illustrated.  The gas sample enters one end  of the distribution
tube illustrated in the upper portion of Figure 5.  The  1.59 cm I.D.  stain-
less steel distribution tube contains 33 holes evenly spaced in a  row in the
positions indicated.  After leaving the distribution tube through  these  holes,
the gas flows around the distribution tube, through the  0.6 mm throat, and
through the cavity in the sample cell perpendicular to the monitoring beam.
The cavity in the sample cell is shaped to produce a minimum amount of mixing
of the gas before it has passed through the volume traversed by the monitor-
ing beam.  After passing the monitoring beam, the gas continues on through
another throat 2.3 mm high and enters the collection tube in the lower
portion of the sample cell.  The gas leaves one end of the collection tube
and is either exhausted or passed on to other instruments for  further
analysis.  This type of gas flow produces very little turbulence in the
volume traversed by the monitoring beam.  At a high gas-flow rate  of  200
liters per minute, no additional noise on the detector signal  attributable
to turbulence can be observed.

     The function of the distribution tube is to distribute the gas flow
uniformly along the length of the sample cell.  Ideally, portions  of  a small
volume of gas that enter the end of the distribution tube at a given  time
should pass through the holes near one end at the same time that other
portions of the same volume element pass through the holes near the middle or
near the other end.  The mixing and the dwell-time in the distribution tube
are reduced to negligible amounts by a tapered piece of  rod placed inside
the distribution tube, as illustrated in Figure 5.  The  O.D. of the rod  is
approximately 0.1 mm less than the I.D. of the distribution tube;  the rod is
tapered so that the cross-sectional area of the gas volume in  the  distribu-
tion tube decreases in the direction of gas flow.

     A few tests were performed with smoke to determine  the efficiency of the
distribution tube and the mixing inside the sample cell.  The  ends of  the
sample cell were replaced with transparent plexiglass so that  a laser beam
could be passed through the length of the cell at different locations.   Light
scattered by smoke within the sample cell was observed through-the plexiglass
end plates.  While air was being pumped through the sample cell at a  typical
flow rate, smoke was injected into the stream of flowing gas just  ahead  of
the inlet to the distribution tube.   The smoke could be  started or stopped
abruptly.  As a test of the distribution tube, the laser beam  was  directed
along the cell approximately 1 cm below the throat near  the tube,  and  the air
flow was adjusted to between 100 and 150 liters per minute.  When  the  smoke
flow, was stopped abruptly,  the smoke disappeared from the laser beam  in  less
than 0.1 sees,  with no distinguishable difference between the  times it dis-
appeared from each end.   This result indicates that the  distribution  tube per-
forms its function quite well.

     The holes in the collection tube are larger than those in the distribu-
tion tube in order to reduce the pressure drop and maintain the absolute
pressure in the sampling volume only slightly above the  outside air pressure.
                                     16

-------
            DISTRIBUTION TUBE
                TAPERED ROD
URETHANE INSULATION (2.5 em THICK)—
                                                                  33 HOLES (1.5 mn DIA) EVEKLY SPACED
                                                                 DISTRIBUTION TUBE (1.6 ca I.D.)
                                                               — GAS INLET
                                                                THROAT  0 .6 mm
                                                               — URETHANE INSULATION (1.3 cm THICK)
                                                                  FIN
                                                                  VOLUME TRAVERSED
                                                                  BY MONITORING BEAM
                                                                    STAINLESS STEEL PLATES (1.2 nm THICK)
                                                                     1.5 cm
                                                                 THROAT 2.3
                                                                COLLECTION TUBE (2.2 cm I.D.)
                                                               GAS OUTLET
                                                                 HOLES  (2.6 mm dla) EVENLY SPACED
            Figure 5.  Sectional view of the sample cell. The left-hand portion shows
            the tapered rod in the distribution tube.  The main body of the cell is
            approximately 31  cm high and 43 cm in the direction parallel to the
            monitoring beam.
                                               17

-------
The spacings between the holes were varied along  the  length  of  the  collection
tube so that the gas flow is nearly uniform  throughout  the length of  the cell.
If the holes were spaced uniformly, more  gas would move across  the  end  of the
cell from which the gas leaves the collection  tube than across  the  opposite
end.  The proper spacings of the holes were  determined  experimentally by
observing light scattered by smoke from a laser beam  located between  the
sampling volume and the 2.3 mm throat. Various combinations  Of  holes  in the
collection tube were covered and the smoke observed.  The distribution  of the
open holes used in the final design produced a flow-pattern  so  that the
smoke disappeared from both ends of the cell at very  nearly  the same  time
after the smoke injected into the airstream had been  stopped.

     The fin placed just upstream from the monitoring beam divides  the  gas
stream into two equal parts.  This fin was not included in the  original
design, but the flow tests performed with smoke indicated that  there  was more
turbulence and mixing within the cell than had been anticipated.  Addition  of
the fin reduced the apparent mixing and therefore reduced the gas turn-over
time.

     The main body of the sample cell is  made  from two  nearly symmetrical
pieces of aluminum that have been milled  out and  placed together as indicated
in Figure 5.  The two pieces are held together rigidly  by 12 bolts.   Silicone
rubber provides a seal at the junction of the  two main  side  pieces  and  at the
junctions of these pieces with the end plates.  A vacuum-tight  seal is  not
required because the gas sample is flushed continuously and  the pressure is
very near atmospheric.  The pressure in the  sample cell is slightly above
atmospheric when the gas is pumped through in  the normal manner as  discussed
in Section 6.  If the gas is drawn through by  a pump  on the  downstream  side of
the sample cell, the pressure is slightly below atmospheric.  The difference
between the sample pressure and atmospheric  pressure  depends on the flow rate
and on the amount of constriction by the  valves,  filter, etc. between the
sample cell and the atmosphere.  The end  plates for the sample  cell,  not shown
in the figures, support the multiple-pass optical system and the windows
through which the monitoring beam enters  and leaves the sample  cell.  Gas fit-
tings also mounted to the end plates connect to the distribution tube and the
collection tube so that these two tubes can  be removed  easily and cleaned,  if
necessary.

     The aluminum body of the cell is anodized to prevent corrosion.  Sensi-
tivity tests performed with the cell when it was  first  assembled indicated
that the anodized surface adsorbed SC>2 from dilute mixtures  at  such a rapid
rate that it would produce erroneous readings  unless  the gas flow was excep-
tionally high.  In addition, some of the  862 that had been adsorbed previously
would desorb from the walls and contaminate "zero-gases" or  mixtures  of lower
concentration as they were flushed through the cell.  After  testing several
different coatings for the anodized aluminim to reduce  adsorption while main-
taining good resistance to corrosion, we  coated the inside surface  with
KEL-F 800.  The solid material was dissolved in a solvent and sprayed on the
anodized surface and heated for a total of about  5 hours at  different tempera-
tures up to 425°F.   Tests indicate that the rate  of adsorption  of S02 on this
material is slow enough that it can be accounted  for by slowly  flushing the
sample through the cell.
                                     18

-------
     The two thin stainless steel plates  on either side of  the  volume  tra-
versed by the monitoring beam were  added  after  the aluminum sides  were made
and some preliminary tests were performed.   These  plates reduce the width of
the gas stream, decreasing the amount  of  mixing and,  in turn, reducing the
gas turnover time.  The surfaces of these plates were roughened by sand-
blasting to reduce the specular reflection  of light from the surfaces  on
which "spill-over" of the monitoring beam is incident.   Extra light from the
oversize monitoring beam is incident on these surfaces  at a grazing angle,
causing the reflection to be quite  efficient.   The reflected light is  a
possible source of error because it does  not travel the same distance  in the
sample as does the main beam.  In addition, this reflected  light makes align-
ment of the multiple-pass optics confusing. No Kel-F coating was  put  on
these stainless steel plates because it would form a glossy surface and
reduce the effect of sandblasting.   No coating  on  stainless steel  is required
because the adsorption of S02 on this  metal is  negligible.

CELL HEATING

     Condensation of water in the sample  cell is avoided by heating the cell
to  any desired temperature up to approximately  60°C.   Electrical heating
tapes, not shown in Figure 5, are bonded to  the  sides of the cell and provide
the heat.  Heating wires attached to the  inlet  and outlet gas lines provide
additional heat to these parts.  A  temperature  sensor embedded  in  the  side
of  the cell near the monitoring beam provides the  input signal  to  a control-
ler that varies the power to the heater as  required to  maintain a  constant
temperature.  The temperature to which the  cell is regulated is varied by
adjusting a small potentiometer on  the temperature controller.   When the
instrument was delivered to EPA, the temperature controller was adjusted to
approximately 55°C, which is high enough  to avoid  condensation  under most
operating conditions  even if some  parts  of the cell are a  few  degrees cooler
than the part near the sensor.  A transformer in the heater circuit limits the
maximum voltage across the heaters  to  36  volts. During typical operation, the
duty cycle of the heater is between 20% and 40%.  Approximately 2  hours are
required to bring the cell from room temperature to 55°C.

     Two thermocouples make it convenient to monitor temperatures  with a high
input-impedance voltmeter.  One of  the thermocouples is embedded in the
aluminum side of the cell adjacent  to  the sensor for the controller.   This
makes it possible to determine the  control  temperature and  to monitor  the
performance of the controller.  The other thermocouple extends  through the
side of the cell wall into the gas  at  a point just downstream from the moni-
toring beam.  This thermocouple monitors  the gas temperature, which may
fluctuate and differ by a few degrees  centigrade from the cell  walls.  The two
thermocouples are made of iron and  constantan wires;  the emf produced  is
0.052 mv for each °C difference between the sensing junction and the refer-
ence junction of the thermocouple.
                                     19

-------
     Sheets of urethane insulation approximately 1.5 cm thick  insulate  the
main cell body.  Four stainless steel posts, each approximately  1.2  cm  long
and 1.2 cm in diameter support the cell body and insulate it thermally  from
the remainder of the optics assembly.  A solid aluminum partition  (see
Figure 4) separates the sample cell from the grating assembly  and  the elec-
tronics.  During typical operation, the grating assembly and the electronics
are less than 5°C above room temperature.
OPTICAL COMPONENTS OF SAMPLE CELL

     Figure 6  illustrates  the  optical portions of  the  sample cell.  A side
view of the multiple-pass  optical  system is shown  in the upper panel.
Mirrors Cl, C2,  and C3 are nearly  flush with the inside surfaces of  the end-
plates so  that  they interfere  a minimum amount with the gas flow.  An image
of  the deuterium arc source is formed beside mirror C2 and just  inside
lens L2, which  also serves as  a window.  Spherical  mirrors Cl, C2, and  C3
have a common  radius of curvature  that is  equal to  the distance  from mirror
C2  to Cl or C3.  Images are formed on mirror C2 as  indicated in  the  lower
panel of Figure 6 when the optics  are adjusted for  12  passes.  The image
pattern is viewed through  the window below mirror C3.  The number  of passes
is  changed by  adjusting mirror C3.  When adjusting  the optics, the tungsten-
iodide bulb is  used as an  energy source because it  emits much  more visible
energy than the arc.  The  centers  of lenses LI and  L2  are separated  by
approximately  30 mm, and the openings in the endplates near these  lenses are
approximately  5 mm in the  long dimension.  Mirrors  Cl  and C2 are each 16 mm
square.  Thus  the cross section of the volume of gas being monitored is
approximately  16 mm by 32  mm.

     The oblong opening in the endplate near lens L2 is overfilled by the
enlarged image  of the deuterium arc source.  Therefore, the images on mirror
C2  have the shape of the oblong opening in the endplate.  Lens L2  forms an
image of lens  LI near mirror Cl.   Lens L3  images mirror C3 near  spherical
mirror X3  (see  Figure 4).  Flat mirrors XI and X2 are  each tilted  approxi-
mately 45° from the vertical.  The combination of these two mirrors  lowers
the beam and directs it horizontally to mirror X3,  which forms an  image on
the entrance slit of the grating assembly.  The image  formed at  lens L3 is
rotated 90° by  mirrors XI  and X3 with the  result that  the long dimension of
the image at the entrance  slit is  parallel to the length of the  slit.   The
entrance slit,  0.3 mm wide by  1.5 mm high, is over-filled in width and
slightly under-filled in height..
                                      20

-------
                 I END FLUX OF SAMPLE CELL
          END PLATE OF SAMPLE CELL
    MONITORING BUM
  MOUNT FOR MIRROR c2
MIRROR XI
                               LENS L2

                                   SILICONS RUBBER SEAL
                                        MIRROR ci-
                              — MIRROR 02
                                       MIRROR 03-
                                   SILICORI ROBBER SEAL


                               LEHSL3
                                10
                        MIRROR C2-
CD CD-
                                             ENTRANCE IMtCE
                                   CD CD

                                   CD CD
                                        CD  «
                                                                       MODNTING POST FOR MIRROR Cl
                                                                   	LEVER ARM FOR MIRROR Cl
                                                                       LETER ARM FOR MIRROR C3
                                                                       MOWTISG POST FOR MIRROR C3
                                                                     WINDCU
                                                                               -PRISM
                                           EXIT IMftGE
        Figure 6.   Multiple-pass  optical system.   The
                      cross-sectional views illustrate  the
                      manner  in  which the optical  compo-
                      nents are  assembled.   The image
                      pattern on mirror C2  is  shown in  the
                      lower panel.
                                     21

-------
                                  SECTION 6

                         PUMP  AND MANIFOLD ASSEMBLIES
      The  pump  assembly  and  the  manifold  assembly are designed to provide the
 user  with a wide  variety of sample flow  conditions.   Sample flow rates can
 be  varied easily  from less  than 5  1pm to more than 160 1pm.  Flow rates are
 adjusted  with  the manifold  assembly and  by  the choice of one of two pumps.
 During  normal  operation, sample gas is drawn into the system to a pump that
 forces  it through the manifold  system and sample cell.  The sample cell
 pressure  is slightly  more than  1 atm.  By re-arranging some of the gas-
 handling  components,  the pumps  can be made  to draw the sample gas through the
 sample  cell before  the  gas  passes  through the pump or the manifold assembly.
 This  latter procedure may be useful if it is found that the pump or the mani-
 fold  assembly  changes the concentration  of  S02-   All of the following discus-
 sion  pertains  to  normal operation  during which the sample gas passes through
 the pump  before it  reaches  the  sample cell.

 PUMP  ASSEMBLY

      The  pump  assembly with  two  pumps mounted  side-by-side  is  represented
 schematically  in  Figure  7.   Sample  flow  rates  as  high  as  approximately  160  1pm
 can be  achieved with  the high-capacity pump  if all of  the gas  lines  are short
 and have  sufficiently large  diameters.   Approximately  80  1pm flow rate  can  be
 achieved  with  the smaller pump.

      A  three-way  electric switch is mounted  on the pump  assembly  so  that
 either  the high-capacity pump or the  low-capacity pump can  be  turned on.  The
 operator  must  check several  valve positions  and  settings  on both  the pump
 assembly  and the  manifold assembly before turning on either pump.  Two  three-
 way ball  valves mounted  on  the plexiglass front  of the pump assembly are
 used  to switch the gas  lines.  If the high-capacity pump  is to be used,  these
 two three-way valves must be rotated  to  the  high-flow  position so the gas can
 pass  from the pump to the manifold  assembly.  Valves on  the manifold assembly
 must  also be opened to pass  the  sample gas.   Similarly,  the valves must be
 rotated to the low-flow  position before  the  low-capacity  pump  is  turned on.
 The three-way valves and  the electrical  switch for the pumps are  color-coded.

      Connected to the line leaving each  pump are  a pressure gauge, a pressure
 switch and a pressure-relief valve.  The pressure gauge  is  useful in monitor-
 ing the output pressure  of the pump which must be kept below 10 psig to avoid
damaging the pump.  The  adjustable pressure  switch automatically  turns  off  the
electrical power  to the pump when the pressure exceeds approximately 9  psig.
The pressure-relief valve opens  at about the same pressure  and serves  as an
additional safety feature.


                                      22

-------
     Compression of the sample gas by  the  pumps  heats the gas,  making addi-
tional heating of the outlet lines unnecessary.   The amount of  heating
increases with increasing pressure in  the  outlet line.   When the high-capacity
pump is operating with 8 psig pressure,  the outlet line may reach a tempera-
ture as high as 85°C.


MANIFOLD ASSEMBLY

     The manifold assembly,  shown schematically in Figure 8, makes it possible
to  adjust  the  rate  of  flow  of  the sample gas to the desired level.  Low-flow
rates  are  read from a  small flow meter with a full-scale reading of 80 stan-
dard cfh  (38 1pm).   A  larger flow meter with a full-scale reading of 6 cfm
 (170 1pm)  measures  higher flow rates.   A 3-way ball valve is used to direct
the gas through the appropriate meter.  The drop in pressure from the outlet
of  the manifold to  the sample  cell can be determined by comparing the pres-
sures  indicated by  the gauges  at these two locations.

     Both  the  flow-adjust valve and  the bypass valve are used simultaneously
to  adjust  flow rates.   Opening the bypass valve prevents the pressure from
building up to a high  level when the flow-adjust valve is adjusted for a  low
flow rate.  Approximate pump pressures can be determined easily from the
gauge  near the inlet to the manifold.

     Calibration gas is flushed through the manifold to the sample cell by
opening a  toggle valve in the  calibration-gas line.   A standard fitting on the
end of the calibration-gas  line makes  it easy to connect it to  a container of
gas mixture with a  known concentration of S02.  The flow-adjust valve is
closed when calibration gas is introduced into the sample cell.

     The  flow  meters,  gauges,  valves and gas lines on the manifold assembly
are heated to  approximately 55°C to  avoid condensation of t^O.   A 36 -volt
 transformer provides electrical power, at a safe voltage, for the heating wire
wrapped around all  of  the components that are in direct contact with the
sample gas.  A contact switch  attached to the gas line opens, stopping the
electric  current, when the  temperature reaches approximately 55°C.   A small
light  indicates when electrical power  is being supplied to the  heating wire.
The switch-point can be adjusted to  other temperatures.   All of the heated
components are packed  in insulation  and enclosed in a metal box.   Handles for
the valves extend through the  front  panel of the box.  Glass windows on the
front  panel make it possible to see  the scales on the gauges and flow meters.
                                       23

-------
                     TO MANIFOLD ASSEMBLY
                      ->~
             H
                        3-WAY BALL VALVE
PRESSURE
   GAUGE
     i
(3 ujn PORE SIZE)
         A
                    IN ~> • FILTER
RELIEF
VALVE
               PRESSURE
                SWITCH
             RELIEF
             VALVE
                                                  PRESSURE
                                                   GAUGE
                       3-WAY  BALL VALVE
                                                    ll
                                   PRESSURED!
                                    SWITCH   !
       /   HIGH
       I    FLOW

           PUMP
                      /   LOW
                         FLOW
                      V   PUMP
      Figure   7.  Flow diagram of pump assembly.
                            24

-------
  TO S02 MONITOR-<~~
                                               /   LOW FLOW METER
HIGH FLOW METER  \
                                                         CALIBRATION GAS
                                                          TOGGLE VALVE
             FROM S02 MONITOR
        DISCHARGE   - < ' <-
                                                               FLOW ADJUST VALVE
PUMP PRESSURE
                                                        A
     PUMP
                                     BYPASS
           Figure   8.   Flow diagram of manifold assembly.
                                     25

-------
                                 SECTION 7

                    SPECTROSCOPIC PRINCIPLES OF DETECTION
     Figure 9 shows a spectrum of the ultraviolet S02 band that is employed
by the instrument.  This band is used rather than the well-known 4 ym band,
which is much weaker and therefore requires longer optical paths in order
to achieve the required sensitivity.  It is noted that the part of the band
shown in Figure 9 shows some relatively sharp structure, but little structure
appears on the short wavelength side of the bandj  Each of these structural
features that stand out in the region near 3000 A consists of many over-
lapping lines.  For samples near 1 atm pressure, these many lines are broad-
ened so that they overlap and there is little remaining structure that is not
resolved in the spectrum shown in Figure 9.  At low sample pressures, many of
these individual lines could be resolved with a spectrometer having very high
resolving power.

     In order to simplify the problem of handling the sample gas, it is
highly desirable that the sample be maintained near 1 atm pressure.  There-
fore, the very fine structure in the spectrum cannot be exploited for de-
tection purposes.  It is necessary to make use of the relatively coarse
structure that is observed in Figure 9.  Figure 10 shows a portion of the
spectrum near 3000A with the wavelength scale expanded so that the lines can
be seen more clearly.  The instrument makes use of this relatively coarse
structure by comparing the transmission through the gas in the 3 intervals
labeled with A to the transmission through the two intervals labeled with B.
The A intervals include very strong absorption by the S02, whereas the B
intervals contain weaker absorption.  The two B intervals are the same width
and are each 50% wider than each of the A intervals so that the total width
of the spectral intervals passed by the B intervals is the same as that by
the A intervals.

     The average transmissions through intervals A and B are compared by
use of the spinning reticle illustrated in Figure 11.  After passing through
the grid, the energy of the selected wavelength intervals is incident on the
spinning reticle.  The reticle is bonded around a cylindrical piece of alumi-
num tubing which has 12 equally spaced slots cut along its long axis and is
secured to the shaft of a motor, which rotates at 1800 rpm.  The reticle has
6 rows of 3 slits and 6 rows of 2 slits cut in it.  The three slits are
aligned directly behind the three grid openings that correspond to slit
intervals A.   The two slits are aligned directly behind the grid openings that
pass intervals B.  The openings in the reticle are oversized with respect to
the grid openings so that this spinning component will not contribute signif-
icantly to instabilities, noise, or zero drift in the output signal.  The
                                      26

-------
t
w
u
M
H
                        100% Transmittance
  2600
2800                3000


   WAVELENGTHS (Angstroms)
3200
   Figure 9 . Spectrum of S02 from 2600 to 3300 Angstroms.   The sample

              consisted of 0.1 atm of pure S02 in a 0.4 cm  cell.
                                    27

-------
      f
      ta
      I
      CO
             •n^Bnot
             /
             100% Transmittance
        2950
      3000

WAVELENGTH  (Angstroms)
3050
Figure 10 .  Partial spectrum of  S02  band showing spectral slits A and  B.
                                    28

-------
to
VO

o
Reference Pick-Up
Motor Shaft 	 >JI /
r°^^^^ P _ 30 Hz Tuned
	 BHI SJ fc Preamplifier l^llf T*

I }
: IXDlIlL 	 Spinning
• ^QnHB Reticle


yu i I
/ l 1
Grid


i 3^6 Hz Gain Controlled
"] Amplifier Amplifier
|




Adjustable iO Hz 30 Hz Tuned
Time Constant * Synchronous * Signal
Amplifier Demodulator Amplifier
t

r v
Fixed Variable
Output Output
Adjustable
-* Phase 	
Shifter

Adjustable
Shifter








i r

nouu j. au ion 360 Hz
». caj-ancc w Demodulator
Assembly


AGC
Control
Amplifier

Calibrate Range
Pot Switch
' '
9^,,,.,,. 	 — . 	 ,.._ .„,„, 	
1 r
Digital
Output
                        Figure 11 .  Block diagram of signal-processing electronics.

-------
energy passing through the openings in the spinning reticle  travels  radially
to the photocathode of the side-looking photomultiplier, which has an  S-5
spectral response.  The photomultiplier is mounted with its  axis  coincident
with the axis of the rotating chopper-alternator.  As the reticle spins, the
beam is chopped at 360 Hz and the energy incident on the photomultiplier
alternates at 30 Hz between spectral intervals A and spectral intervals B.
The relative widths of the openings corresponding to spectral intervals A
and B is adjusted so that the same amount of energy is incident on the photo-
multiplier during both halves of the cycle of the spinning reticle as  long as
there is no SC>2, or other absorbing gas, in the sample cell.  In  practice, it
is not possible to maintain exactly this equality between the two intervals,
but the slight  imbalance  is accounted for electronically by a method
described in Section 8.  After some experimentation, it was  determined that
the cathode of the photomultiplier has a non-uniform sensitivity  and thus
the response of the photomultiplier to photons incident on different regions
of the cathode is not the same.  The imbalance created by this non-uniform
sensitivity is also accounted for electronically by the method described in
Section 8.

     A 30 Hz component of the detector signal results when S0£ is added to
the sample cell, and the size of this signal depends on the  source bright-
ness, amount of dirt on the windows, and on any other parameters  that  might
change the total amount of energy incident on the slit of the grating  assem-
bly.  In order to avoid the problem of the detector output signal being
dependent upon these parameters, the instrument is made to measure a quantity
that is proportional to the fractional modulation of the beam, rather  than to
the total change in energy striking the detector.  This is accomplished by
chopping the beam at 360 Hz.  When S02 is present in the sample cell,  the
360 Hz carrier signal is modulated at 30 Hz.  The carrier signal  is main-
tained at a constant level by an automatic gain control circuit.  The  elec-
tronics described in Section 8 process the detector signal in such a way as
to produce a dc output that is proportional to (Tg - TA)/(TB + TA.) .  This
quantity is proportional to the fractional modulation of the beam and  is
related directly to the concentration of SC>2 in the sample cell of fixed
pathlength.  The changes in source brightness, detector responsivity,  and
electronic gain are automatically accounted for so that the  "span" calibra-
tion remains constant.
                                     30

-------
                                  SECTION  8

                  ELECTRONIC PROCESSING OF DETECTOR  SIGNAL


     Figure 11 is a block diagram of  the  electronics  that process the signal
from the detector and produce a dc voltage that  is proportional to the con-
centration of S02-  The relationship  between  the output voltage and the gas
concentration is determined by calibrating the instrument with samples of
known concentration.  The electronics consist basically of two demodulators
in series.  The first demodulator operates at 360 Hz, the carrier frequency
determined by the chopper.  The average dc component  in the output from the
first demodulator is proportional to  the  amount  of radiant energy chopped by
the spinning reticle.  The 30 Hz  component of the signal output of the first
demodulator is proportional to the modulation of the  beam at this frequency.
This 30 Hz component then passes  through  a series of  gain controls and
amplifiers to the second demodulator, which is synchronous and receives its
reference signal from a reference pick-up mounted near the rotating grid
chopper.  The output of the second demodulator is then proportional to the
30 Hz modulation, and thus to the concentration  of S02 in the sample cell.

     The amplified component of the detector  signal at the carrier frequency,
360 Hz, is defined to be Vc.  Correspondingly, the component at the modula-
tion frequency, 30 Hz, is defined as  Va.  It  is  desirable for the instrument
output to be proportional to the  fractional modulation of the carrier signal.
This fractional modulation is proportional to Va/Vc,  which we define as V,
and it is desired that this relationship  be independent of source brightness,
detector sensitivity, etc.  An automatic  gain control (AGC) circuit maintains
Vc at a constant value.  If, for  example, the signal  from the detector
decreases because of dirt on the  windows  or a decrease in source brightness,
the gain of the gain-controlled amplifier increases  to maintain Vc constant.
The increase in gain also increases the amplification of both the 360 Hz
carrier signal and the 30 Hz modulation signal by the same factor.  Therefore,
V is kept directly proportional  to the output Va and maintains a constant
relationship with the gas concentration.  The AGC circuit can account for
changes in the 360 Hz component of the signal by approximately a factor of 4.

     The reference pick-up for the 30 Hz  synchronous  demodulator consists of
a light-emitting diode (LED) and  a small  phototransistor.  The phototransistor
senses the light reflected from the metallic  tape wrapped around the outside
of the spinning reticle.

     A zero-balance assembly electronically accounts  for any imbalance
between the two beams passing through spectral intervals A and spectral
intervals B when there is no absorbing gas in the sample cell.  This
                                     31

-------
electronic assembly would not be required if it were practical to obtain a
perfect optical balance between the two sets of slit intervals and if the
cathode of the photomultiplier had a uniform sensitivity.  Approximately a
+ 10% variation in the transmittance through either of the sets of intervals
can be accounted for by this electronic zero-balance assembly.  The assembly
changes the amplification ahead of the 360 Hz demodulator during the two
halves of the cycle of the spinning reticle.  The signal from the 30 Hz
reference pick-up switches the zero-balance assembly between the two different
gains with the proper phase.  The difference in the gains between the two
halves is adjustable by the "zero" potentiometer mounted on the front panel
of the instrument.

     Five ranges of sensitivity are provided; 5, 10, 20, 50, and 100 ppm.
The gain in each range can be adjusted independently over a factor of
approximately 4.  A six-position switch mounted on the front hinged panel of
the optics assembly makes it convenient to change ranges.  When the switch is
in the off position, the output of the 360 Hz synchronous demodulator is
shorted to ground.  Because of optical saturation, it is necessary to reduce
the number of passes to 2 or 4 passes in order to operate in the 0 - 1000 ppm
range.  All ranges must be recalibrated with appropriate span gases for
operation in the 0-1000 ppm range.

     A four-position switch mounted on the main electronics card makes it
possible to use the analog panel meter to monitor other quantities in addi-
tion to the signal output.  With the switch in one position, the meter
indicates the level of the reference signal; with the switch in another
position, the meter reads the level of the 360 Hz signal from the photo-
multiplier before the automatic gain control circuit; with the switch in
still another position, the meter indicates the signal input level.  Thus it
is possible to check to see that both the reference signal and the carrier
level are within the correct operating range.

     The digital panel meter directly indicates the 862 concentration in the
sample gas.  The relationship between the analog meter indication and the
concentration, of course, depends on the range to which the instrument is set.
The digital panel-meter is connected to an output jack so that the digital
output can be recorded or transmitted over a line to another instrument.  The
size of the analog voltage output that corresponds to full-scale reading of
the instrument can be varied continuously from less than 1 mV to 10 V.  Thus,
this analog output can be recorded easily with virtually any strip-chart
recorder with a high input impedance.
                                     32

-------
                                 SECTION  9

                        CALIBRATION AND PERFORMANCE
SPAN CALIBRATION

     Each of the five sensitivity ranges  (5,  10,  20,  50,  and  100 ppm full-
scale) has been calibrated with  the  sample  cell controlled at 55°C.  The
calibration depends on the cell  temperature because the instrument response
is approximately proportional to the number density of S02 molecules, re-
gardless of temperature.  If the S02 concentration in ppm remains constant
and the pressure is one atm, an  increase  in temperature results in a pro-
portional decrease in the number density  of S02 molecules in  the monitoring
beam.  Thus, the instrument response decreases and the potentiometers used
to vary the gains of the electronic circuits must  be adjusted  so the meters
will indicate the proper concentration.   In most  cases, a small correction
to the calibration for all five  ranges  can be made by adjusting a single
potentiometer that changes the gains for  all  five ranges by the same factor.

     At a fixed temperature, the output signal is essentially proportional
to S02 concentration for concentrations less  than 30  ppm.  At higher con-
centrations, the response becomes slightly  non-linear, necessitating small
corrections.  The meters and the voltages on  the  output jacks are directly
proportional to the output signal; no circuits have been added to account
for the non-linearity.  Table 1  contains  the  information required to correct
for the non-linear response when using the  50 ppm and 100 ppm full-scale
ranges.  No correction is required for the  5, 10, and 20 ppm  full-scale
ranges.
                                      33

-------
              TABLE  1.   CORRECTIONS TO CALIBRATION TO
                         ACCOUNT FOR NON-LINEAR OUTPUT
Meter Reading        Correction         Corrected Reading
    (ppm)               (%)	(ppm)	
                 50 ppm Full-Scale Range

   ^30                  0                Same as meter

     40                 +1                     40.4

     50                 +2                     51.0


                 100 ppm Full-Scale Range

   § 30                 -4                0.96 x meter reading

     40                 -3                     38.8

     50                 -2                     49.0

     60                 -1                     59.4

     70                  0                     70.0

     80                 +1                     80.8

     90                 +2                     91.8

    100                 +3                    103.
     Errors due to the non-linear relationship between SC>2 concentration and
output signal are negligible for the 5, 10 and 20 ppm full-scale ranges, and
for concentrations lower than 30 ppm measured on the 50 ppm full-scale range.
The approximate corrections required for the 50 ppm and 100 ppm full-scale
ranges are given above.  The gain for the 100 ppm full-scale range has been
adjusted so that it is nearly correct at 70 ppm.  Thus the maximum error due
to non-linearity on the 100 ppm full-scale range is approximately 3% for any
sample that would normally be measured on this range.

     The same corrections apply whether the values are read from the analog
panel meter, the digital panel meter, or one of the output jacks.
                                  34

-------
ADSORPTION OF S02 ON CELL WALLS

     Sulfur dioxide adsorbs  readily to the walls of containers.   The adsorp-
tion process may take  place  for  several minutes before the amount adsorbed on
the walls comes to an  equilibrium.   The equilibrium amount depends on the
concentration of the S02 vapor in contact with the walls.   The S02 monitor
senses only the concentration of S02 in the vapor phase;  therefore sizeable
errors can be made if  adsorption is ignored or not accounted for properly.
In order to minimize the problem due to adsorption, we have coated the inside
surface of the sample  cell with  Kel-F 800, an inert plastic material that is
noted for low adsorptivity.   Nevertheless, some care in sampling is still
required because of adsorption in the gas lines and the small amount that
occurs in the sample cell.

     The simplest method of  accounting for the adsorption  is to  flush the
gas sample continuously through  the sample cell so that the S02  that adsorbs
on the walls is replenished  before the vapor concentration is depleted
significantly.  "Zero" gas being used when adjusting the zero-setting should
also be flushed to carry away S02 that comes off the cell  wall after having
been adsorbed from previous  samples.

     A few simple tests were performed to determine the rates of adsorption
and the rate of sample flow  required to avoid significant  errors in the
measurements.  The first test involved introducing a sample of 3.9 ppm S02
concentration into the sample cell after it had been flushed with N2 long
enough that there was  essentially no S02 adsorbed on the walls.   The 3.9  ppm
mixture was flushed rapidly  through the cell for 30 sec, a period long enough
for the S02 vapor concentration  in the cell to reach its maximum value.   The
gas flow was then stopped  abruptly, and the S02 vapor concentration was moni-
tored for a few minutes. At one minute after the flow was stopped, the vapor
concentration had decreased  because of adsorption to 92% of its  original
value.  The concentration  continued decreasing and was down to 75% of the
original value after 5 minutes.   A similar test was performed with a 76 ppm
mixture; the rate of percentage  decrease was essentially the same as for  the
leaner mixture.

     Another simple test was performed by introducing a 76 ppm mixture into
the "clean" sample cell at different'flow rates.  A constant flow rate of 5
liters per minute was  adequate to maintain the S02 vapor concentration within
less than 1% of the maximum  concentration that would be measured with much
higher flow rates.  The 5  1pm flow rate represents the minimum that can be
measured reliably with the  low-capacity flowmeter on the manifold assembly.
Lower flow rates may be adequate, but if they are used, checks should be
made for possible errors due to  adsorption.
                                     35

-------
NOISE LEVEL AND ZERO-DRIFT

     The dominant source of noise is the photoelectron noise generated in
the photomultiplier.  An oscilloscope trace of the photomultiplier output
shows no apparent noise during the periods of each cycle when the spinning
reticle blocks the uv energy from the photocathode.  The noise level, at
30 Hz frequency, on the photomultiplier output was found to be approximately
proportional to the square-root of the uv energy level on the photocathode;
this relationship is expected when photoelectron noise is dominant.  The
output of the electronics that process the photomultiplier signal is pro-
portional to the ratio of the 30 Hz component to the 360 Hz component.  The
360 Hz component is proportional to the uv energy level on the detector;
therefore the 30 Hz noise level on the instrument output is inversely pro-
portional to the square-root of the incident uv energy level.  It follows
that for a given S02 concentration, the signal-to-noise ratio is proportional
to the square-root of the incident uv energy level.

     The noise level on the 5 ppm full-scale range was measured over a
5-minute interval with no S02 in the sample cell, the electronic time
constant at 1 sec, and the deuterium arc source operating at 0.7 amps, the
maximum current obtainable with the power supply.  The output signal was
recorded on a strip-chart recorder, and the digital output meter was read
once each 5 sec.  The peak-to-peak noise level corresponded to approximately
0.2 ppm of S02, and the rms noise level as determined from the 60 digital
readings was 0.049 ppm.  As expected, the noise level decreases when the
electronic time constant is increased.

     The zero-setting is quite stable, particularly after the instrument has
been "warmed-up."  During one test period of 30 minutes, the zero-setting,
as averaged over a 1-minute interval, drifted less than 0.05 ppm.  The
amount of zero-drift in practical use will depend mostly on the temperature
stability of the air surrounding the instrument.
                                     36

-------
SAMPLE TURN-OVER TIME

     Table 2 sumarizes  the  results  of  a series of tests  performed  to  deter-
mine the time response  of the  instrument to an abrupt change in the S02
concentration of the gas entering the  sample cell.   Time resolution,  rather
than the actual response time,  is the  quantity of most interest.   In  a
practical application,  there is a delay between the time of  a sudden  change
in S02 concentration at the exhaust pipe of an automobile and the  time that
the gas reaches the sample  cell where  the change is sensed.   This  delay
depends on the gas lines and on the flow rate and will be different for
different systems.  This delay can be  measured and accounted for so that the
proper time relationship can be determined between the observed concentration
change and any changes  in the  engine operation.

     An important factor in the design of the sample cell and gas-handling
system has been the need to maintain good time resolution.   Consider  the case
in which the S0£ concentration at the  inlet to the filter on the pump assem-
bly changes abruptly from level A to a different level B.  After a certain
delay, the instrument reading  starts deviating from level A  and goes  through
a continous change to level B.   This period of transition of the instrument
reading is a measure of the time resolution, which is a  function of gas flow
rate, mixing in the sample  cell, and electronic  response time.  The very
gradual change in the instrument reading as it approaches level B  makes it
impractical to measure  the  complete transition time.   In the laboratory tests,
we measured times required  for 70% and 90% of the change.  In every case the
period started when there was  the first distinct change  in the deflection of
the recorder that monitored the output signal.

     Test No. 1 was performed  to determine the response  of the recorder to a
sudden change in its input  signal.   The recorder signal  was  started at approx-
imately 80% of full-scale;  the input jack was unplugged  suddenly,  and the re-
corder response was measured.   This measurement, of course,  does not  involve
the main part of the S02 monitor.  As  in all of  the other tests, the  recorder
chart speed was 1 inch  per  second.   The estimated error  in the measured times
is approximately 0.05 sec for  periods  less than 1 sec, and somewhat more for
longer periods.  The values tabulated  represent  averages of  several indepen-
dent readings.

     Test No. 2 provides information on the response of  the  electronics,
including the photomultiplier, the signal processing electronics and  the
recorder.  The results  of this test, or the previous one, have nothing to
do with the mixing of the gas  in the sample cell.  A small mask was held so
that it covered a portion of one of the slits of set A in the grid.   This
produced an output signal that corresponded to approximately 17 ppm of S02.
While recording the output  signal,  we  abruptly removed the mask.   Approxi-
mately 0.35 sec was required for the recorder to make 70% of the change; 0.45
sec was required for a  90%  change.   These times represent the absolute
minimum that we would expect to measure when changing the S02 concentration
if no mixing of the sample  took place.  The electronic time  constant  was 0.1
sec for this test and for all  tests involving the main electronics.
                                      37

-------
  TABLE 2.   RESPONSE TIMES FOR STEP CHANGE IN S02
             CONCENTRATION AT VARIOUS SAMPLE FLOW RATES
Test
No.
1
2

3
4
5
6
7
8
9
Change Made
or
Flow Rate in 1pm
Unplugged Recorder
Input Cable
Uncovered Portion
of Grid
160a
160b
130b
85b
38b
19b
9.5b
Time for
90% Change
(Sec)
0.29
0.48

0.74
0.74
0.78
1.18
1.25
1.96
3.68
Time for
70% Change
(Sec)
0.25
0.35

0.45
0.48
0.55
0.73
0.88
1.43
2.52
Electronic time constant is 0.1 sec for all tests.

aSample introduced at filter on pump assembly

 Sample introduced at distribution tube of sample cell
                             38

-------
     It is apparent from  the  results  of tests  No.  1 and No.  2  that  the
recorder adds significantly to  the  measured response times.  Somewhat more
accurate data on the sample turnover  time could be obtained  with  a  faster
recorder; however, the  lack of  a rapid recorder response probably added no
more than 0.1-0.2 sec to  the  times  measured when the gas concentration was
changed.

     Test No. 3 was performed by injecting a slow, steady flow of pure S02
into the gas stream just  as it  entered the filter  on the pump  assembly.  The
injection rate of the S02 was adjusted to produce  a concentration in the line
of approximately 17 ppm of S02.   While recording the output  signal, the flow
of pure S02 was stopped abruptly in such a way that there was  no  slow diffu-
sion of S02 from the injection  needle after the flow was stopped.   The 70%
and 90% turnover times  of 0.45  and  0.74 sec indicate that mixing  in the gas
line and sample cell prevents a gas "front" from moving sharply across the
monitoring beam.

     Test No. 4 was performed by the same method as Test No. 3, except that
the pure S02 was injected into  the  line just ahead of the distribution tube
so that it did not travel through the pump and manifold.  The  results of
Tests  3 and 4 are essentially in agreement, indicating that  most  of the
breaking-up of the gas  front  occurs in the sample  cell, not  in the  pump
or manifold assemblies.

     Tests 5 through 9  were identical to Test  No.  4, except  that  the sample
flow rates were varied.  As expected, the turnover times decrease with in-
creasing flow rates, but  not  in the same proportions.  There is apparently
less breaking-up of  the gas front at low flow  rates.  For example,  in re-
ducing the flow rate by a factor of 9, from 85 1pm to 9.5 1pm, the  90%
turnover time increases by a  factor of 3.68/1.18 =3.1.  It  is apparent that
little is gained by  operating at flow rates above  130 1pm.

     If a much faster recorder  were used, the  minimum 90% turnover  time would
probably be between  0.6 and 0.7 sec.   This period  is somewhat  higher than
the 0.5 sec design objective.
                                       39

-------
INTERFERENCE BY OTHER EXHAUST GASES

     Interference by automotive exhaust gases other  than  S02  is  negligible
at the concentrations these gases normally occur  in  the exhaust.   No measurable
interference was found for the following hydrocarbons  tested  individually at
Aeronutronic:  acetylene, methane, butane, propane,  ethylene  and ethane.   Each
sample consisted of a mixture of a single hydrocarbon  gas in  N2  at 1 atm total
pressure.  The data on hydrocarbon interference and  other data obtained at
Aeronutronic are summarized in Table 3 along with still other  data obtained  at
EPA by EPA scientists.

     Many of the interference data were obtained  with  a small sample cell
temporarily placed in the monitoring beam while the  main  multiple-pass  sample
cell was by-passed.  This improved the signal-to-noise ratio  because of the
reduction in reflective losses in the multiple-pass  optics.   The short  sample
cells also enabled us to use smaller amounts of gas  at higher concentrations
than would be used in the large, multiple-pass cell.   The mixtures of higher
concentration are easier to mix accurately,  and  there is less potential
error due to adsorption on the walls of the sample cell.   Table  3 gives the
cell lengths and the gas concentrations along with the equivalent concentra-
tions for the multiple-pass cell.
     Ammonia  (NHo), nitric oxide  (NO), and nitrogen dioxide  (M^)  each  inter-
 fere slightly, but the amount of  interference produced by the  low  concentra-
 tions  in  automotive exhaust is negligible.  Because of the detector noise and
 the very  slight  interference, the discrimination ratios could  not  be measured
 accurately.   Nitric oxide absorbs quite strongly in the spectral band employed
 by the instrument, but there is very little correlation between the structures
 in the S02  spectrum and the N02 spectrum.  Thus, there is very little inter-
 ference .

     When the 100% C02 test gas first entered the sampling volume, the  output
 signal indicated an 862 concentration of approximately 0. 2 ppm, but this sig-
 nal vanished  as  the C02 completely displaced the N2 that was originally in
 the sample  cell.  The temporary signal was apparently caused by a  non-uniform
 distribution  of  the C02 in the cell.  Differences in the indices of refraction
 of C02 and  ^ probably produce small deflections in the monitoring beam when
 the gases are both present in the cell and are not mixed.  Absorption by  C02
 is completely negligible in the spectral interval passed by the instrument.

     A small  signal was also observed some of the times when an ^0 mixture
 first  entered the sample cell that was previously filled with  N2.  The  size
 of this transient signal was not  reproducible,  but it never corresponded to
 more than 0.2 ppm of S02.  As in  the cases when C02 was introduced into the
 cell,  the transient signal disappeared when the cell became completely  filled
 with either the  ^0 mixture, or with N2 when the ^0 mixture was being  flushed
 out.   When  the ^0 mixture was in the sample cell, the 360 Hz  component of the
 detector  signal  was typically about 1% lower than it was when  the  cell  was
 filled with No,   The 1% attenuation is more than can be accounted  for by ^0
 vapor  absorption.  This attenuation is not understood, but  it  may  be  due  to
 a change  in the  reflectivity of the mirrors of the multiple-pass optical
 system that results from a layer  of absorbed water on the reflecting  surfaces.
                                    40

-------
     The lower part of Table  3  summarizes  the results  of tests  performed
on two complex mixtures  simulating automobile exhaust  hydrocarbons. These
mixtures did not contain other  common exhaust species  such  as CO, C02,
N0y, or aldehydes  such as  formaldehyde

     The samples were prepared  by injecting either gasoline or  diesel  fuel
into a heated tube while purging with nitrogen,   The vaporized  fuel samples
were collected in  large  Tedlar  bags and the total hydrocarbon concentration
in ppm determined  with a flame  ionization  detector (F.I.D.) hydrocarbon
analyzer,  The overall compositions of the two fuels are given  as a footnote
in Table 3.

     The aromatic  species  in each of these samples would be the most likely
to  cause interference  as they tend to absorb in the U.V.  close  to the  300nm
region which the 862 monitor operates.  The aromatics  in the gasoline  are
mostly benzene, toluene  and other alkyl benzenes.   The diesel fuel aromatics
contain alkyl benzenes,  alkyl naphthalenes, alkyl anthracenes,  alkyl phenanth-
rences and benzothiophenes.

-------
                             TABLE  3.  INTERFERENCE DATA
Test
No
1
2
3
4
5
6
7
8
9 '

10

11

Gas
Acetylene
Methane
Butane
Propane
Ethylene
Ethane
Ammonia
Nitrogen-
Dioxide
Nitric
Oxide
Carbon
Dioxide
Water
Vapor
Concentr-
tion
7o
100
100
100
100
100
100
100
25
40

100

2.5

Path
Length
(cm)
3
3
3
3
3
3
3
0.4
0.4

480

480

Equiv. Con-
centration
(ppm)
6250
6250
6250
6250
6250
6250
6250
208
333

106

25,000

Interfer- Discrimi-
ence Error nation Ratio
(ppm of S02>
N.D.a
N.D.
N.D.
N.D.
N.D.
N.D.
0.02 b
0.32 c
0.01 d

N.D.

N.D.

N.D.
N.D.
N.D.
N.D.
N.D.
N.D.
330,000:
650:
30,000:

N.D.

N.D.







1
1
1





 The following tests on gasoline and diesel fuel were performed at EPA by EPA personnel.
El
E2
E3
E4
Gasoline e
Gasoline
Diesel e
Fuel
Diesel
Fuel
100 ppm
THC
400 ppm
THC
100 ppm
THC
180 ppm
THC
480
480
480
480
100
THC
400
THC
100
THC
180
THC
N.D.
N.D.
N.D.
N.D.
N.D.
N.D.
N.D.
N.D.
(a)  N.D. is the abbreviation for Not Detectable
(b)  Absorptance of the ammonia sample was 0.02
(c)  Absorptance of the N02 sample was     0.027
(d)  Absorptance of the NO sample was      0.04
(e)  The overall compositions of the two fuels were:
                           Gasoline      Diesel Fuel
7o Paraffins
7, Olefins
7» Aromatic s

47.3
7.8
44.9
100.0
82.4
0.4
17.2
100.0
                                       42

-------
TECHNICAL REPORT DATA
(Please read Instructions on the reverse before completing)
"^REPORT NO. ~" I ~ '
EPA-600/2-79-121
4. TITLE AND SUBTITLE









DEVELOPMENT OF AN S00 MONITOR FOR MOBILE SOURCES
7 AUTHOR(S) *•


jp.E. Burch, P.S. Davila. F.J. Gates, and J.D. Pembrook
9. PERFORMING ORGANIZATION NAME AND ADDRESS ~~^
Ford Aerospace and Communications Corporation
Aeronutronic Division
Ford Road
Newport Beach. California


29663



12. SPONSORING AGENCY NAME AND ADDRESS
Environmental Sciences Research Laboratory
- RTP, NC
Office of Research and Development
U. S. Environmental Protection Agency
Research Triangle Park. North Carolina 27711
15. SUPPLEMENTARY NOTES


16. ABSTRACT " 	
An instrument has been designed and built to monitor the
exhaust of mobile sources. A grating assembly desperses
deuterium-arc source and passes five narrow
3. RECIPIENT'S ACCESSION NO.
5. REPORT DATE
Oiily 1979


6. PERFORMING ORGANIZATION CODE


8. PERFORMING ORGANIZATION REPORT NO
U-6392
10. PROGRAM ELEMENT NO.
1AA601B CA-29 (FY-78)
11. CONTRACT/GRANT NO.
68-02-2448


13. TYPE OF REPORT AND PERIOD COVERED
Final 10/76 - 1/78
14. SPONSORING AGENCY CODE


EPA/600/09

concentration of S0» in the
the ultraviolet energy from a
spectral intervals. Three of the intervals
(set A) coincide with strong absorption features in the S02 band near
3000 A; the other
two intervals (set B) coincide with weak absorption. A spinning reticle alternately
transmits energy passing through set A and
put signal proportional to
appropriate electronics.
B to a photomultiplier detector. A dc out-
the concentration of S0? in the sample cell is produced by




A pump and manifold assembly permit operation over a wide range of sample flow rates.
At high flow rates, the 90%
sample turn-over time is approximately 0.7 sec. The sample
cell contains a multiple-pass optical system adjusted to
path length greater than 5
12 passes, giving a sample
m. The rms noise level corresponds to approximately 0.05
ppm of SO. when the electronic time constant is 1 sec.
•
17.
•™— — »— ^ — ^ ________ „ _________«__ _______»__
a- DESCRIPTORS
	 	 	 	 	
* Air Pollution
* Motor vehicles
* Exhaust emissions
* Sulfue dioxide
* Monitors
* Development
»?. DISTRIBUTION STATEMENT 	
RELEASE TO PUBLIC


KEY WORDS AND DOCUMENT ANALYSIS ~~
— ___• •! Ill !!•_• 	 ••






•^•^•••••^•^•••••••••^•i^^B^^^HIIII^HMVBIIIIBIMV



b.lDENTIFIERS/OPEN ENDED TERMS
Correlation-spectrometer










19. SECURITY CLASS (This Report 1
UNCLASSIFIED
20. SECURITY CLASS (This page)
UNCLASSIFIED
e. COSATI Field/Group
13B
13F
21B
07B


21. NO OF PAGES
51
22. PRICE

Form 2220-1 (Rev. 4-77)    PREVIOUS EDITION is OBSOLETE
                                                  43

-------