&EPA
           United States
           Environmental Protection
           Agency
          Environmental Monitoring and Support  EPA 600 4-79-015
          Laboratory          February 1979
          Research Triangle Park NC 27711
           Research and Development
Ozone
Calibration and
Audit by Gas Phase
Titration in Excess
Ozone

Bendix®
Transportable Field
Calibration System,
Models 8861 D and
8861 DA

-------
                RESEARCH REPORTING SERIES

Research reports of the Office of Research and Development, U.S. Environmental
Protection Agency, have been grouped into nine series. These nine broad cate-
gories were established to facilitate further development and application of en-
vironmental technology.  Elimination of traditional grouping  was consciously
planned to foster technology transfer and a maximum interface in related fields.
The nine series are:

      1.  Environmental  Health Effects Research
      2.  Environmental  Protection Technology
      3.  Ecological Research
      4.  Environmental  Monitoring
      5.  Socioeconomic Environmental Studies
      6.  Scientific and Technical Assessment Reports (STAR)
      7.  Interagency Energy-Environment Research and Development
      8.  "Special" Reports
      9.  Miscellaneous Reports

This report has been assigned to the ENVIRONMENTAL MONITORING series.
This series describes research conducted to develop new or improved methods
and instrumentation for the  identification and quantification of environmental
pollutants at the lowest conceivably significant concentrations.  It also includes
studies to determine the ambient concentrations of pollutants in the environment
and/or the variance of pollutants as a function of time or meteorological factors.
This document is available to the public through the National Technical Informa-
tion Service, Springfield, Virginia  22161.

-------
           OZONE CALIBRATION AND AUDIT BY  GAS  PHASE TITRATION
                             IN EXCESS OZONE
(Bendix  Transportable Field Calibration System,  Models 8861D and 8861DA)
                                   by

                  Thomas A.  Lumpkin and Barry E.  Martin
                          Field Studies Section
             Environmental Monitoring and Support Laboratory
                  U.S.  Environmental Protection Agency
              Research  Triangle Park, North Carolina  27711
             ENVIRONMENTAL MONITORING AND SUPPORT LABORATORY
                   OFFICE OF RESEARCH AND DEVELOPMENT
                  U.S.  ENVIRONMENTAL PROTECTION AGENCY
              RESEARCH  TRIANGLE PARK, NORTH CAROLINA  27711

-------
                                  DISCLAIMER










     This report has been reviewed by the Environmental Monitoring and Support




Laboratory, U.S. Environmental Protection Agency, and approved for publication.




Mention of trade names or commercial products does not constitute endorsement




or recommendation for use.
                                      11

-------
                                   ABSTRACT







     Detailed procedures for the dynamic calibration and audit of chemilu-



minescence ozone analyzers are presented.  These procedures were developed and



applied within the Environmental Monitoring Branch, Environmental Monitoring



and Support Laboratory, U.S. Environmental Protection Agency, Research Triangle



Park, North Carolina.  The purpose of the procedures is to aid calibration and



audit personnel in performing calibrations and audits in exactly the same man-



ner with identical calibration systems.






     The calibrations and audits are performed by means of a gas phase titration



technique utilizing the rapid gas phase reaction between nitric oxide and ozone



with excess ozone present.  The nitric oxide is generated by using a cylinder



of nitric oxide in nitrogen that has been standardized against a National Bureau



of Standards Standard Reference Material.  The instrument being calibrated or



audited must have a linear response to ozone.  An ozone concentration is gener-


                                               ®
ated using a dynamic calibration system  (Bendix  Model 8861D or 8861DA) and



is introduced to the analyzer under calibration or audit to obtain an up-scale



response.  A known concentration of nitric oxide is added to excess ozone in



the calibration system and the change in analyzer response noted.  Under spe-



cifically controlled conditions, the decrease in analyzer response is equal



to both the concentration of ozone consumed and the concentration of nitric



oxide added.  The original ozone concentration (before nitric oxide was added)





                                      iii

-------
can then be calculated.  Other calibration concentrations of ozone can be ob-




tained by diluting the original ozone concentration and repeating the gas phase




titration.









     One of the advantages of these procedures is that chemiluminescence ozone




analyzers can be calibrated or audited in the field without the bulky equipment




required for the neutral buffered potassium iodide calibration procedure.  A




second advantage is that more precise results can be obtained.  A standardized




cylinder of nitric oxide replaces the potassium iodide solution, the set of




bubblers, the vacuum system, and the spectrophotometer required for the potas-




sium iodide method.
                                      IV

-------
                       LIST OF ABBREVIATIONS AND SYMBOLS
ABBREVIATIONS



  3
cm



GPT-0



hr



1



min



mol



NBKI



NBS



ppm



psig



s



SRM



URL
— cubic centimeter



— gas phase titration in excess ozone



— hour



- liter



— minute



— mole



— neutral buffered potassium iodide



— National Bureau of Standards



— parts per million



— pounds per square inch gauge



— second



— Standard Reference Material



— upper range limit
SYMBOLS





N2


NO




N02


NO
— nitrogen



— nitric oxide



— nitrogen dioxide



— total oxides of nitrogen



— oxygen
                                     Vll

-------
0.
                                      — ozone

                                      — Input A pressure gauge setting

                                      — Input B pressure gauge setting

                                      — Input C pressure gauge setting

                                      — triiodide ion
 'NO
 NO

V V F2
                                      — cylinder NO concentration, ppm

                                      — flow through 0  generator, cm /min

                                      — NO  flow rate, cm /min

                                      — total dilution air  flows, cm /min

                                      — analyzer recorder response after
                                        addition of NO to system, % chart
Z80' Z40'
                                      — analyzer recorder  response  to 0  ,
                                        %  chart
              ' etc-
 [03]80'  [°3]60' etC'
(PAE)8(),
                , etc.
— adjusted analyzer recorder response
  to 0 , % chart

— 0  concentration at URL of 80%,
  60%, etc., ppm

— percent audit error at URL of 80%,
  60%, etc.
                                       — adjusted analyzer recorder response,
                                         % chart
 U
                                      — unadjusted  analyzer  recorder  response,
                                         %  chart
                                     Vlll

-------
                                ACKNOWLEDGMENTS









     The personnel of the Field Studies Section, Environmental Monitoring




Branch, contributed important input after testing and using the procedures.




Special recognition is due Mr. Kenneth Rehme and Mr. Frederick Smith of the




Monitoring Techniques Evaluation Section, Environmental Monitoring Branch, for




the initial testing and evaluation of the gas phase titration in excess ozone




technique.
                                       IX

-------
                                   SECTION 1









                                 INTRODUCTION









     The Federal Register specifies the neutral buffered potassium iodide




(NBKI) method as the procedure for calibration of designated reference methods




for measurement of photochemical oxidants (1).  This procedure consists of




passing a generated ozone (0 )  concentration through a NBKI solution and then




measuring the triiodide ion (I ) concentration with a calibrated spectrophoto-




meter.  The 0  concentration can be determined from the I  concentration and




the known volume of air passed through the NBKI.  The basic equipment is an O




generator, a dilution air supply, a flowmeter, a bubbler train, a vacuum




system, and a spectrophotometer.  Commercially available calibration systems




usually combine a dilution air supply with an 0  generator.  The remaining




equipment can become difficult to manage and operate on field calibrations,




especially if air travel is involved.  Often, due to lack of proper facilities,




the NBKI procedure cannot be performed as precisely in the field as in the




laboratory.  Also, the spectrophotometer may require recalibration after being




moved from place to place in the field.  Recalibration of the spectrophotometer




involves exact standards and is difficult to accomplish under field conditions.
     This report describes procedures for performing O  calibrations and audits.




These procedures — gas phase titration in excess ozone — utilize a standard

-------
nitric oxide (NO) gas cylinder as the reference standard.  The only other




equipment necessary is a flowmeter and a calibration system capable of sup-




plying specified stable O  concentrations at specified flow conditions.  Al-




though a significant difference exists between these procedures and the NBKI




procedure, the difference is consistent.  Thus the procedure can be referenced




to the NBKI procedure specified in the Federal Register.

-------
                                   SECTION 2







                        CONCLUSIONS AND RECOMMENDATIONS







     Using these procedures, the calibration and audit of 0  analyzers requires



less complicated equipment than with the NBKI procedure.  In addition, more



consistent and more precise results can be obtained in the field.  These pro-



cedures have been used successfully by the Environmental Monitoring Branch on



two previous studies:  the 1975 Summer Ozone Study and the 1976 Washington,



D.C. Oxidant-Hydrocarbon Study.  The procedures are presently being used on



the National Forest Ozone Study which began in 1976.






                                                                    ®
     Although the procedures have been written for a modified Bendix  Model


                 ®
8861D or a Bendix  Model 8861DA Transportable Field Calibration System, any



comparable calibration system can be used by modifying the procedures slightly.



Any different calibration system must be thoroughly tested and shown to be



consistent with the NBKI procedure.

-------
                                   SECTION 3
                                   PRINCIPLE
     Ozone calibration by gas phase titration in excess ozone (GPT-0 ) is




based upon the rapid gas phase reaction between O  and NO in accordance with




the following equation (1,2):
                 NO + O  -»• NO  +0       k = 1.0 x 10  1/mol s
An O  concentration of unknown magnitude is generated in a dynamic calibration




system and sufficient NO of known concentration is added to decrease the O.




concentration by 90-95% of its original value as measured on an uncalibrated




chemiluminescence 0, analyzer.  If the exact NO concentration is known, the




concentration of 0  can be determined and can then be used to calibrate the O




analyzer.  The standardization of the NO cylinder used in this technique is




based on direct comparison with a certified National Bureau of Standards  (NBS)




NO Standard Reference Material  (SRM).  The flow conditions used in the dynamic




calibration system are optimized to ensure the quantitative reaction of NO




with 0  and to minimize the reaction of nitrogen dioxide  (N0?) and O  which




can lead to errors in the calibration procedure.  Erroneous results will also




occur if the analyzer response is nonlinear.  The analyzer linearity is verified




by a dilution technique described in Section 6.

-------
                                   SECTION 4
                                   APPARATUS
      /*v
BENDIX' TRANSPORTABLE FIELD CALIBRATION SYSTEM, MODEL 8861D OR MODEL 8861DA
     Figures 1 and 2 illustrate Bendix  Models 8861D and 8861DA, respectively.
The Model 8861D system should be modified as described below and as shown in
the flow diagram in Figure 3.  In Figure 3, the modified sections are enclosed
in broken lines.  The modifications make the calibration system more effective
and versatile in calibrating and auditing 0  analyzers by these procedures.


Modification of Bendix  Model 8861D


Step 1—
     Replace the 0-15 psig NO pressure gauge with a 0-25 psig gauge.  Dismantle
the O  generator-flow capillary compartment and replace the NO flow capillary
with one that will give a flow rate of about 20 cm /min at 25 psig.  This
modification allows higher concentrations of NO to be obtained from the calibra-
tion system at the specified dilution flow rates.

-------

                          ©
Figure 1.   Modified Bendix  Model 8861D Transportable Field Calibration System.

-------
                 -
Figure 2.   Bendix  Model 8861DA Transportable Field Calibration System.

-------
CLEAN
 AIR

CLEAN-UP
LAMP





r




FILTER


















( 	 '
^ 	 ,


                                                               AIR PUMP
                                           WATER DROP-OUT
Figure 3.  Flow Diagram for modified Bendix  Transportable Field Calibration  System.

-------
Step 2-



     While the compartment is dismantled, remove the orifice fitting that con-



trols the flow through the 0  generator and plug the hole from which the fitting



was removed.  Replace the fitting at the inlet to the 0  generator with the


                                                               ®
orifice fitting.  Connect an 18-in length of 1/8-in o.d. Teflon  tubing to the


                                                                           ®
orifice fitting.  Close the compartment, leaving the free end of the Teflon



tubing outside the compartment.  Connect an additional pressure regulator and



gauge (0-25 psig, preferably similar to original equipment) to the zero air



supply upstream from the existing regulator and gauge that control the dilu-


                                  ®
tion air flow.  Connect the Teflon  tubing to the outlet of the pressure regu-



lator just installed.  This modification allows the flow through the 0_ gener-



ator to be controlled separately by the new regulator and gauge.  The modifica-



tion is needed primarily for calibrations and audits of 0  analyzers by the



GPT-0  procedure.  Although the modification is not essential for calibrations



and audits of NO/NCv/NO  analyzers, the system can still be used to calibrate
                   ^   X


such analyzers.
Step 3-


     Remove the condensing coil from the system.  Cut the electrical wires to


the fan and remove it from the system.  Insulate the ends of the cut wires.


Install a silica gel scrubber  (0.5 1 or larger capacity) in line between the


water dropout and the cleanup lamp.  This modification gives a drier calibra-


tion air than that obtained with the condenser-fan assembly.
                                       11

-------
Step 4-



     Replace the small charcoal-soda lime filter downstream from the surge



tank with one with a capacity of at least 0.5 1.  This will allow the calibra-



tion system to be operated for a longer time between scrubber material changes.
0  ANALYZER
     Any chemiluminescence 0- analyzer that has a rapid and linear response



to 0  can be calibrated by this technique.  Ozone analyzers using other measure-



ment principles can be calibrated if they do not respond to NO or NO».
FLOWMETER
     A bubble flowmeter kit and/or a wet-test meter capable of measuring abso-



lute flow rates between 1 and 5000 cm /min is required.
PRESSURE REGULATOR







     The standard NO cylinder requires a pressure regulator with stainless


                               ®
steel internal parts and Teflon  seats.







SAMPLE MANIFOLD
     A Kjeldahl mixing bulb (approximately 300 cm  volume) with a multiport



glass manifold is recommended.
                                      12

-------
                                   SECTION 5







                                   REAGENTS







STANDARD NO CYLINDER







     The cylinder should contain approximately 100 ppm NO in nitrogen (N ).



The NO content of the cylinder is determined by comparison with a NBS NO SRM.







ZERO AIR





               ®
     The Bendix  Model 8861 Transportable Field Calibration Systems supply air



free of contaminants that would cause a detectable response in the 0  analyzer



or interfere with the 0  calibration.  The zero air supply of the Model 8861D



System should include two scrubbing columns; the first column should contain



indicating silica gel and the second should contain 1/2 soda lime and 1/2



activated charcoal, in that order.  These two scrubbing columns should be



replaced after 8 hr of actual use, or sooner  (if indicated by the silica gel).



The Model 8861DA System should have one scrubbing column containing 1/2 soda



lime and 1/2 activated charcoal, in that order.  This column should be re-



placed after 8 hr of actual use.
                                       13

-------
                                  SECTION  6






                             CALIBRATION PROCEDURE





                                      ®
     The flow conditions in the Bendix  Model 8861 Transportable Field Calibra-



tion Systems are optimized for this procedure and must be duplicated precisely



to ensure the validity of the calibration.
STEP 1
     Record pertinent information about the analyzer being calibrated in the


space provided on the data sheet  (see Appendix A).  If a request for informa


tion does not apply, write "N/A" in the space.  In the margins, record any


pertinent information needed for a particular analyzer but not specifically


requested on the data sheet.  Adjust all analyzer flow rates to the manufac-


turer's specifications.
STEP 2
     Connect the standard NO cylinder to the Bendix  calibration system.  Open


the cylinder valve and let the NO flow rate stabilize.  The pull-to-test valve


for NO should be out while the 0  generator and dilution air flow rates are


measured.  Place the sample manifold on the outlet of the calibration system.



                                      15

-------
(See Figure 4 for the arrangement of the calibration apparatus.)  Switch



capillaries 1, 2, 3, and 4 of Input C  (dilution air) to the "OFF" position.



Set the air flow through the 0  generator to 150-200 cm /min by adjusting the



Input A gauge pressure.  Measure the flow using a bubble flowmeter attached



to the outlet of the calibration system.  (The sample manifold must be dis-



connected each time a flow measurement is made at the outlet of the system.)



Record the 0  generator air flow (F ) and the Input A pressure gauge setting
            j                      G


(P ) on the data sheet.
STEP 3
     Set the total air flow to approximately 2500 cm /min by adjusting the



Input C gauge pressure (any combination of capillaries 1, 2, 3, and 4 can be



used).   Using a bubble flowmeter or wet-test meter, measure the total air flow



at the outlet of the calibration system.  Record the total air flow  (F ), Input



C pressure gauge setting  (P ), and capillaries used on the data sheet.
STEP 4
     Advance the O  analyzer recorder chart a few inches from the last ambient



air trace and allow the analyzer under calibration to sample its internal zero



air (if applicable) until a stable response is obtained.  Then allow the ana-



lyzer to sample calibration zero air until a stable response is obtained.



Record the unadjusted recorder response  (Z ) for the analyzer's zero and for



calibration zero air.  Make the proper zero adjustment to the analyzer using
                                      16

-------
 STAINLESS
   STEEL
 REGULATOR
                          CALIBRATION SYSTEM

                          (BENDIX ,  MODEL 886ll> OR
                            MODEL 886 IDA)
GLASS MIXING BULB AND MANIFOLD
                           FLOW
                                           BUBBLE FLOWMETER
                                          (OR WET TEST METER)
            i—I n r~< r
      STANDARD
      NO CYLINDER
                                                          OZONE ANALYZER
Figure 4.  Flow scheme for calibrations and audits by gas phase titration in excess ozone.

-------
the calibration zero air as the reference.  Record the adjusted recorder



response (Z )  and adjusted zero setting on the data sheet.
STEP 5
     Switch the 0  generator to the appropriate range and adjust the set point



dial to provide an 0  concentration of approximately 80% of the upper range



limit  (URL) as measured on the 0  analyzer.  Record the analyzer response  (Ion)
                                •j                                           oU


and the 0  generator set point on the data sheet.
STEP 6
     Add the NO flow by pushing the pull-to-test valve in and adjust the Input



B pressure regulator until the O  analyzer response has decreased by 90-95%



of its original value.  For example, if I   = 85% chart and Z  = 5% chart, the



NO flow should be adjusted to yield an analyzer response of 9-13% chart.  Re-



cord the analyzer response (I) after it has stabilized.
STEP 7
     Measure the NO flow using a bubble flowmeter attached to the pull-to-test



valve.  Record the NO flow  (F  ), cylinder NO concentration  (C  ), and Input B
                             NO                               NO


pressure gauge setting  (P ) on the data sheet.
                         B
                                      18

-------
STEP 8
     Calculate the exact NO concentration from:
                                [NO] -                                (Eq.
                                       NO    0
where     [NO] = NO concentration, ppm
          F ^ = NO flow, cm /min
           NO
          C   = cylinder NO concentration, ppm
           NO
           F  = total air flow, cm /min
Record the calculations on the data sheet.
STEP 9
     Calculate the 0  concentration from:
                                    I   - Z
                            [°3]80 = I   - I  *  tNO]                  (Eq-  2)
where     [0-J0n = 80% URL 0  concentration, ppm
           3 oU            3
           [NO] = NO concentration, ppm
            I0/. = original O_ analyzer response, % chart
             bU             3
              1=0  analyzer response after addition of NO,  % chart
             Z  = adjusted zero air recorder response,  % chart
                                      19

-------
Calculate the response to which the recorder should be adjusted as  follows:
                                    [°3]80

                          J8o(A) =	URT—+ZA                 (E^  3)
where    I   (A) = adjusted recorder response, % chart
          oU


            URL = upper range limit, ppm



Record the calculations on the data sheet.
STEP 10
     With the NO flow removed, the 0  analyzer response should return  to  its



original value  (Ior.).  If the response does not return to within  ± 1%  chart
                 80


of I  , recheck I by adding NO to the system again.   If I is exactly the  same
    oU


as before, remove the NO and let the 0  response return to  its new value.



Record this new response as I   and recalculate  [O  ]   and  I   (A).  If I  is
                             oO                   3 oU      oO


not the same as before, measure the NO flow again and recalculate [NO],  [O_]o
                                                                           j  o


and Iori(A).  After the O  response has stabilized up-scale, record this new
     80                 j


value as I  .  Adjust the span control until the analyzer recorder gives  the
          oU


desired response as .calculated in Equation 3.  Record the adjusted recorder



response [I  (A)] and adjusted span setting on the  data sheet.
STEP 11
     Adjust the Input C gauge pressure to give a  total air  flow  of  approximately



3300 cm /min.  Using a bubble flowmeter or wet-test meter,  measure  the  total
                                       20

-------
air flow at the outlet of the calibration system.  Record the total air  flow




(F ) , Input C pressure gauge setting  (P  ) , and the capillaries used on the  data
  J.                                    L-



sheet.  Calculate the diluted O_ concentration at 60% URL  ([0,]c.) from:
                               J                             3 DU
                              [°3]60=  [°3]80Xi                     (E<5'  4)
where     [O.]rr. = 60% URL 0, concentration, ppm
           3 60            3


          [0 ]   = 80% URL 0  concentration, ppm
           380            3


             F  = original total air flow, cm /min



             F  = total air flow after dilution, cm /min



Record the calculations and the recorder response  (I-.) on the data  sheet.
                                                    oO
STEP 12
     Readjust the Input C gauge pressure to give the original  total  air  flow



 (F ).  Repeat Steps 5 through 10, substituting 40% URL  for 80% URL,  I    for



I  , and  [O ]   for [0 ]  .  Disregard the reference in Step 9 to  adjusted
 80        3 40       3 oO


recorder response.  Make no further adjustment to the analyzer span  control.



Note;  In Step 6, the 0  response should be decreased by  90-95% of its new
                       3                                               	


value.  For example, if I   = 45% chart and Z  = 5% chart, the NO  flow should
                         ^t \J                  f\


be adjusted to yield an analyzer response of 7-9% chart.
STEP 13
     Adjust the Input C gauge pressure to give a total air  flow of approximately



5000 cm /min.  Using a bubble flowmeter or wet-test meter,  measure the  total



                                       21

-------
air flow at the outlet of the calibration system.  Record the total air flow



(F ) ,  Input C pressure gauge setting  (P ), and the capillaries used on the
  ^                                    V-*


data sheet.  Calculate the diluted 0  concentration  ([0  ]   ) from:








                             [°3]20=  [°3]40X?7                    (B*  5)
where     [0 ]   = 20% URL 0  concentration, ppm



          [0 ] 0 = 40% URL 0  concentration, ppm



             F  = original total air flow, cm /min



             F  = total air flow after dilution, cm /min



Record the calculations and the recorder response  (I5n) on the data sheet.
STEP 14:  RECORDING RESULTS ON DATA SHEET
     a.  Record the 0, concentrations generated and the analyzer responses



obtained at these concentrations.  Also record the zero and span settings



before and after adjustments were made.
     b.  Plot the analyzer response  (y axis) versus 0  concentration  (x axis)



 (see Figure Al).  The analyzer response should be linear within  ±  1%.
     c.  Summarize any problems encountered during the calibration.   If  this



was a field calibration, name persons from local agencies present during the



calibration.
                                       22

-------
     d.  Record on the 0  analyzer recorder chart the name of the station




calibrated, the date, information for each trace on the chart, and the time the




calibration was begun.  If possible, remove the calibration segment of the




recorder chart and attach it to the data sheet.  Advance the recorder chart,




synchronize the time, and date it.  If a portion of the recorder chart is re-




moved, explain on remaining chart that a calibration took place.  Do not destroy




any ambient air data while cutting the chart.
                                       23

-------
                                   SECTION 7







                                AUDIT PROCEDURE







     If the Model 8861D Calibration System is used, it should be modified as


                                                          ®
described in Section 4.  The flow conditions in the Bendix  Model 8861 Trans-



portable Field Calibration Systems are optimized for this procedure and must



be duplicated precisely to ensure the validity of the audit.
STEP 1
     Record pertinent information about the analyzer being audited in the space



provided on the data sheet  (see Appendix B).  If a request for information does



not apply, write "N/A" in the space.  In the margins, record any pertinent in-



formation needed for a particular analyzer but not specifically requested on



the data sheet.  Do not make any adjustments to the analyzer.
STEP 2
     Connect the NO cylinder to the Bendix  Calibration System.  Open the



cylinder valve and let the NO flow rate stabilize.  The pull-to-test valve for



NO should be out while the 0  generator and dilution air flow rates are mea-



sured.  Place the sample manifold on the outlet of the calibration system.
                                       25

-------
(See Figure 4 for the arrangement of the calibration apparatus.)  Switch



capillaries 1, 2, 3, and 4 of Input C  (dilution air) to the "OFF" position.



Set the air flow through the 0  generator to 150-200 cm /min by adjusting the



Input A gauge pressure.  Measure the flow using a bubble flowmeter attached



to the outlet of the calibration system.  (The sample manifold must be dis-



connected each time a flow measurement is made at the outlet of the system.)



Record the 0  generator air flow (F ) and the Input A pressure gauge setting
            3                      Cj


(P ) on the data sheet.
STEP 3
     Set the total air flow to approximately 2500 cm /min by adjusting the




Input C gauge pressure (any combination of capillaries 1, 2, 3, and 4 can be




used).  Using a bubble flowmeter or wet-test meter, measure the total air flow




at the outlet of the calibration system.  Record the total air flow  (F ), In-




put C pressure gauge setting  (P ), and capillaries used on the data sheet.
                               \*
STEP 4
     Advance the O  analyzer recorder chart a few inches from the last ambient



air trace and allow the analyzer under audit to sample zero air from the calibra-



tion system for about 30 min or until a stable response is obtained.  Record



the unadjusted recorder response (Z ) for audit zero air.
                                       26

-------
STEP 5
     Switch the 0, generator to the appropriate range and adjust the set point



dial to provide an 0  concentration of approximately 80% of URL as measured



on the 0_ analyzer.  Record the analyzer response  (Ior.) and the O_ generator
        j                                           oU           j


set point on the data sheet.
STEP 6
     Add the NO flow and adjust the Input B pressure regulator until the 0



analyzer response has been decreased by 90-95% of its original value.  For



example, if I   = 85% chart and Z  = 5% chart, the NO flow should be adjusted
             OU                  U


to yield an analyzer response of 9-13% chart.  Record the analyzer response



(I) after it has stabilized.
STEP 7
     Measure the NO flow using a bubble flowmeter attached to the pull-to-test



valve.  Record the NO flow  (F  ), cylinder NO concentration  (C  ), and Input B
                             NO                               NO


pressure gauge setting  (P ) on the data sheet.
                         B
STEP 8
     Calculate the exact NO concentration from:

                                      27

-------
where     [NO] = NO concentration, ppm
          F _ = NO flow, cm /min
           NO
          C   = cylinder NO concentration, ppm
           F  = total air flow, cm /min
Record the calculations on the data sheet.
STEP 9
     Calculate the 0. concentration from:
                                    I   ~ Z
                            [°3]80 = i80 - j" x  [NO]                  (E(5'  2)
                                     80
where     [0-]on = 80% URL 0_ concentration, ppm
           -3 oU            J
           [NO] =.NO concentration, ppm
            I   = original 0  analyzer response,  % chart
             ou             j
              I = O  analyzer response after addition of NO,  %  chart
             Z  = unadjusted zero air recorder response, % chart
Record the calculations on the data sheet.
STEP 10
     With the NO flow removed, the 0  analyzer response  should  return to  its
original value  (Ion) .  If the response does not return to within  ±  1% of  I.,.,
                 oU                                                        80
recheck I by adding NO to the system again.   If I is exactly  the  same as  before,
remove the NO and let the 0  response return  to its new  value.  Record this new
                                      28

-------
response as I   and recalculate [0 ]   .   If I is not the same as before, mea-
             oO                   3 oU


sure the NO flow again and recalculate  [NO] and  [0 ]  .  After the O  response
                                                  3 80              3


has stabilized up-scale, record this new value as I0_.  Calculate and record
                                                   oO


the percent audit error (PAE) from:
                                  URL  (I   - Z )


                        (PAE)80 = 	[0
where     (PAE)0l_ = percent audit error at.80% URL
              oU


             URL = full scale range



             I   = original O  analyzer response, % chart
              oO             3


           [0 ]   = 80% URL O  concentration, ppm
            3 oU            3


              Z  = unadjusted zero air response, % chart
STEP 11
     Adjust the set point dial on the 0  generator to give an 0  concentration



of approximately 50% of the URL and repeat Steps 5 through 10, substituting



40% URL for 80% URL, I^Q for IQQ, [0^ 4Q for  [0^^, and  (PAE)4Q for  (PAE)8Q.
STEP 12:  RECORDING RESULTS ON DATA SHEET
     a.  Record the O  concentrations generated and the analyzer responses



and PAE values obtained at these concentrations.
                                       29

-------
     b.   Plot the analyzer response (y axis) versus O  concentration  (x axis)




(see Figure Bl).   The audit limits shown are ± 2% of full scale for zero and




± 15% difference between actual and observed values for up-scale readings.









     c.   Summarize any problems encountered during the audit.  If this was a




field audit, name persons from local agencies present during the audit.









     d.   Record on the 0  analyzer recorder chart the name of the station




audited, the date, information for each trace on the chart, and the time the




audit was begun.   If possible, remove the audit segment of the recorder chart




and attach it to the data sheet.  Advance the recorder chart, synchronize the




time, and date it.  If a portion of the chart is removed, explain on the re-




maining chart that an audit took place.  Do not destroy any ambient air data




while cutting the chart.
                                       30

-------
                                  REFERENCES
1.    Hodgeson, J. A., R. E. Baumgardner, B. E. Martin, and K. A. Rehme.  Stoi-
     chiometry in the Neutral lodometric Procedure for Ozone by Gas Phase
     Titration with Nitric Oxide.  Anal. Chem., 43(8):1123-1126, 1971.

2.    Rehme, K. A., B. E. Martin, and J. A. Hodgeson.  Tentative Method for
     the Calibration of Nitric Oxide, Nitrogen Dioxide, and Ozone Analyzers
     by Gas Phase Titration.  EPA-R2-73-246, U.S. Environmental Protection
     Agency, Research Triangle Park, North Carolina,  1974.

3.    Rehme, K. A.  Application of Gas Phase Titration in the Calibration of
     Nitric Oxide, Nitrogen Dioxide and Ozone Analyzers.  In: Calibration
     in Air Monitoring, ASTM STP 598, American Society for Testing and
     Materials, 1976.  pp. 198-209.
                                      31

-------
1.
                                  APPENDIX A
                      SAMPLE GPT-O  CALIBRATION DATA SHEET
STATION
CALIBRATION PERSONNEL
ADDRESS
DATE
Analyzer Calibrated
Ethylene Cylinder: Mfgr.
Ethylene Flow Rate
Sample Flow Rate
Zero Base Line
S/N
S/N Pressure psig
cm /min
3. .
cm /nun
% chart
     Unadjusted Zero Setting

     Unadjusted Span Setting
     Full Scale Range Selector Switch Position

     Time Constant Selector Switch Position 	

     Mode Selector Switch Position 	

     Calibration System 	

     NO Cylinder:  Mfgr. 	

     Room Temperature 	
                     S/N
S/N
Pressure 	psig  Cone.
ppm
 °C  Barometric Pressure
                          in Hg
     NOTE:  Refer to the "Ozone Calibration by Gas Phase Titration in Excess
            Ozone" procedure.
                                       33

-------
2.    0_ Generator Flow:   F
      j


         Vol.  Meas.
         FG=
)cra3/(
_cm      Time



     )min =
                                      mm
     Input A Pressure Gauge Setting:  P  =
cm /min
                                  % gauge
3.   Total Air Flow:  F,
             Time (min)
    Avg.  Time (min)        Vol. Meas. (cm )
         1)



         2)



         3)
                      )cm /(
    )min =
     Input C Pressure Gauge Setting:  P  =
                                       L^
                cm /min



               	% gauge
     Capillaries Used:   1234   (Circle)
4.   Zero Air Data
                                             Analyzer
                                Calibration
     Unadjusted Recorder Response:
     Adjusted Recorder Response:
                                                                         % chart
     Adjusted Zero Setting:
                                                                         % chart
5.   80% URL Data
     O  Analyzer Recorder Response:  I   =
      3                               80


     O  Generator Setting:  	
                                   % chart
                                       34

-------
6.    0  Analyzer Recorder Response  (with  NO present):   I = 	% chart
7.   NO Flow:  F^
                NO


             Time  (min)           Avg.  Time  (min)        Vol. Meas. (cm )



         1)  	           	       	



         2)  	



         3)
         F _ =  (       )cm /(       )min =                cm /min
          NO                              -


     Cylinder NO Concentration:   C   = _ ppm



     Input B Pressure  Gauge Setting:   P  = _ % gauge
                                        B   -
            FNO X CNO    (        ) (       )    (       )

                         - -  = -  • - ppm
9.    [0 , =           x  [HO]  ,  - 1 -  x ,    ,  ,            = _ ppn,

       *     80                 v      -     ;              v    ;
10.  Span Data



     Unadjusted Recorder  Response:   I   = 	% chart
                                      oU   	


     Adjusted Recorder  Response:   Iori(A)  =                           % chart
                                    ou      ^^^^^^^—^^—^—^^—^^—^^—^^-~—


     Adjusted Span Setting: 	
                                       35

-------
11.  60% URL Data




     New Total Air Flow:  F




             Time (min)




         1)  	




         2)  	




         3)
                        )cm3/(
            Avg.  Time (min)
                )min =
     Input C Pressure Gauge Setting:  P  =



     Capillaries Used:     123
                                                   Vol.  Meas. (cm )
                            (Circle)
                                                        cm /min
                                       % gauge
     t03]60
                                ppm
     0. Analyzer Recorder Response:  I
      3
                               % chart
12.  40% URL Data



     0  Analyzer Recorder Response:  I
                                      40
     0  Generator Setting:
                           % chart
     0  Analyzer Recorder Response  (with NO present):  I =
     NO Flow:  F
                NO


             Time (min)
                                          % chart
            Avg.  Time (min)
                                                   Vol. Meas. (cm )
         1)



         2)



         3)
         FNO=
)cm3/(
                            )min =




Cylinder NO Concentration:  C   =
     Input B Pressure Gauge Setting:  P  =
                                       B
cm /min
                                                 ppm
                               % gauge
                                       36

-------
Input C Pressure Gauge Setting:




Capillaries Used:     1     2
                                      P  =
     [NO]  =
F   x C
 NO    NO


F   + F
 NO    0
                                         % gauge
                                                   (Circle)
                                               ppm
                  - Z
     [°3]40
                                                        ppm
13.   20% URL Data




     New Total Air Flow:  F,
                           <:



             Time (min)




         1)  	




         2)  	




         3)
         F2 =
        )cm3/(
                      Avg. Time  (min)
                                                   Vol. Meas.  (cm  )
                           )min =
     Input C Pressure Gauge Setting:  P  =



     Capillaries Used:     123
 cm /min



	% gauge
                                       (Circle)
      [°3]20
                      ) x
     0  Analyzer Recorder Response:  I   =
                                                     ppm
                                         % chart
14.  Results



     a.  03 Concentration, ppm
          [03]Q =  0.00
          [°3]20=




          [°3]40=




          [°3]60=




          [°3]80=
                           Analyzer Response,  % chart
                                       37

-------
                          Before Adjustment      .      After Adjustment




    Zero Setting          	.	            	




    Span Setting          	            	




b.  (See Figure Al)




c.  (Remarks)
                                  38

-------
cc
<
X
o
   100-
90-
   80-
o

cr
UJ
Q.
70-
S  60~
z
o
°-  50-
co  ^^ '
LJ
ir
cr  40-
N
V
   30-
o
N
O
   20-
     o-


u 1
0
0
0.05
0.10
0.10
0.20
0,15
0.30
0.20
0.40
0.25
0.50
                       OZONE CONCENTRATION, ppm
Figure Al.  Calibration curve — analyzer response versus 0  concentration.
                                39

-------
1.    STATION

     ADDRESS
                                  APPENDIX B
                         SAMPLE GPT-0  AUDIT DATA SHEET
      AUDITOR

      DATE
     Analyzer Audited
     Ethylene Cylinder:  Mfgr.

     Ethylene Flow Rate 	

     Sample Flow Rate 	

     Zero Base Line
       S/N

       S/N
                        Pressure
psig
               cm /min
              _cm /min

               % chart
     Unadjusted Zero Setting

     Unadjusted Span Setting
     Full Scale Range Selector Switch Position

     Time Constant Selector Switch Position 	

     Mode Selector Switch Position

     Calibration System 	

     NO Cylinder:  Mfgr. 	

     Room Temperature 	
     S/N
S/N
         	 Pressure 	psig  Cone.

C  Barometric Pressure 	in Hg
ppm
     NOTE:   This is an audit.  Do not make any adjustments to the analyzer.
            (Refer to the "Ozone Audit by Gas Phase Titration in Excess Ozone"
            Procedure.)
                                      41

-------
2.
0_ Generator Flow:  F
 3                   G
Vol. Meas. cm Time
F^, = ( )cm /( )min =
it A Pressure Gauge Setting: P =
min
cm /min
% gauge
3.   Total Air Flow:  F
                       0
             Time (min)
                             Avg. Time  (min)
                         Vol. Meas. (cm )
         1)



         2)



         3)
         Fo =
               )cm3/(
)min =
     Input C Pressure Gauge Setting:  P  =
                                       \^
 cm /min



	% gauge
     Capillaries Used:
                                               (Circle)
4.   Unadjusted Recorder Response:  Z  =
                                               %  chart  (audit  zero  air)
5.   80% URL 0  Data



     0. Analyzer Recorder Response:  I   =
      3                               o(J
                                                 %  chart
     0- Generator Setting:
6.   0  Analyzer Recorder Response  (with NO present):   I =
                                                                 % chart
                                       42

-------
7.   NO Flow:   FNQ
             Time  (min)           Avg.  Time (min)       Vol. Meas. (cm )
         1)



         2)



         3)
                       3                3
         F   =  (      )cm /(     )min = cm /min



     Cylinder NO  Concentration:  C   = 	ppm



     Input B  Pressure Gauge Setting:  P  = 	% gauge
                                      g   	






           FNO X CNO   (     ) (     )   (      )
8.   [NO]  =  M"   N" = ]	4	f = +	f =  	ppm

             NO   0    V            '   ^      '






9-   [03]Qo = i^r*  [NO] H	~-	fx(     > =  (  (  )(  )   }  =—K»
               ou
Recorder Response: Ior> = % chart
URL
(rAE)80
(I80-V „„ « ,
'°3'80 ' <
( - )
)
                                                      - 100 =
11.  40% URL 0  Data




     0  Analyzer Recorder Response:  I   = 	% chart



     O  Generator Setting: 	
     O  Analyzer Recorder Response (with NO present):  I = 	% chart



     NO Flow:   FNQ




            Time  (min)           Avg.  Time (min)       Vol. Meas. (cm )
         1)



         2)



         3)
         F   =  (      )cm /(     )min = 	cm /min




                                     43

-------
     Cylinder NO Concentration:   C   = _ ppm



     Input B Pressure Gauge Setting:  P  = _ % gauge
                                       B



     .   ,    FNO X CNO   ()()()


     11)01  =             - -  =  -  -
     Recorder Response:  I   = _ % chart





               URL II40 - V
12.   Results




     a.  03 Concentration, ppm     Analyzer Response, % Chart   PAE
         [03]Q =    0.000
                                   i40-
                                   i80=
     b.  (See Figure Bl) .




     c.  (Remarks)
                                      44

-------

—
a:
<
T.
O
—
Z
LU
U
cr
1 1
-%
j_
i
LU
Z
o
a.
C/)
LU

-------
                                   TECHNICAL REPORT DATA
                            (Please read Instructions on the reverse before completing)
  REPORT NO.
  EPA  600/4-79-015
                                                           3. RECIPIENT'S ACCESSION-NO.
4. TITLE AND SUBTITLE
  Ozone Calibration and Audit by Gas  Phase Titration  In
  Excess Ozone   Bendix Transportable Field Calibration
 JLvstem. Models RSfiin and 8861 DA	
                                                           5. REPORT DATE
                                                             February 1979
             6. PERFORMING ORGANIZATION CODE
7. AUTHOR(S)
  Thomas A.  Lumpkin
  Barry E.  Martin
                                                           8. PERFORMING ORGANIZATION REPORT NO.
9. PERFORMING ORGANIZATION NAME AND ADDRESS
  Field Studies  Section
  Environmental  Monitoring and Support  Laboratory
  U. S. Environmental  Protection Agency
  Research Triangle Park, North Carolina  27711
             10. PROGRAM ELEMENT NO.

                1AD606
             11. CONTRACT/GRANT NO.
12. SPONSORING AGENCY NAME AND ADDRESS
  Environmental  Monitoring and Support  Laboratory/RTP
  Office of  Research and Development
  U. S. Environmental  Protection Agency
  Research Triangle Park, N. C.  27711
                                                           13. TYPE OF REPORT AND PERIOD COVERED
             14. SPONSORING AGENCY CODE
                 EPA/600/08
15. SUPPLEMENTARY NOTES
16. ABSTRACT
     Detailed  procedures for the dynamic  calibration and audit  of chemiluminescence
ozone analyzers  are presented.  The  calibrations and audits are performed by means
of a gas phase titration technique utilizing the rapid gas phase reaction between
nitric oxide and ozone with excess ozone  present.

     The purpose of this report is to  aid calibration personnel  in performing
calibrations and audits in exactly the same manner with identical  calibration systems.
One of the advantages of the procedures is that chemiluminescence ozone analyzers
can be calibrated or audited in the  field without the bulky equipment required for
the neutral buffered potassium iodide  calibration procedure.  A second advantage is
that more precise results can be obtained.
17.
                                KEY WORDS AND DOCUMENT ANALYSIS
                  DESCRIPTORS
b.IDENTIFIERS/OPEN ENDED TERMS  C. COSATI Field/Croup
   air pollution
   calibration
   gaseous pollutants
   ozone
   gas phase  titration
  43 F
  68 A
18. DISTRIBUTION STATEMENT

   Release to  Public
19. SECURITY CLASS (This Report)
  Unclassified
21. NO. OF PAGES
 56
                                              20. SECURITY CLASS (Thispage)
                                                Unclassified
                                                                         22. PRICE
EPA Form 2220-1 (9-73)

-------