INNOVATIVE APPROACHES TO DATA
         VALIDATION
 UNITED STATES ENVIRONMENTAL
      PROTECTION AGENCY
          REGION
          JUNE 1995

-------
           INNOVATIVE  APPROACHES FOR
    VALIDATION OF ORGANIC AND INORGANIC DATA-
         STANDARD OPERATING PROCEDURES
                   JUNE.1995
      U.S. .ENVIRONMENTAL PROTECTION AGENCY
                   REGION III
ANALYTICAL SERVICES AND QUALITY ASSURANCE BRANCH
         201 "DEFENSE HIGHWAY, SUITE 200

-------
                                 NOTICE

The procedures set forth herein are intended as guidance.  They do not constitute
rulemaking by the Agency and may  not be relied on to create a substantive or
procedural right.enforceable  by any other person.  The Government may take
action that is at variance with the procedures in this manual.

-------
                               INTRODUCTION
There are five levels of data validation within this manual approach. The levels consist of three
(3) organic review procedures (M-l, M-2, M-3) and two inorganic review procedures (IM-1,
IM-2).  (The term "level" is perhaps misleading because it implies succession, which is not the
case with this or any other level presented in this document. The "levels " are independent of
each other.) M-3 and IM-2 review procedures represent full validation as described in the EPA
National Functional Guidelines for Data "Review. Should the intended data use dictate review
by the protocols presented in the National Functional Guidelines for Data Review (IM-2 and M-
3) refer to the most recent version of Region m Modifications to the National Functional
Guidelines.  The remaining data review levels are described in detail in this document  The
SOPs describing how to apply these levels to a data-package are in Appendices A through D.
All procedures require(full CLP or^ C^ equivalent data .package, deliverable^. -\
         - •ป •     .          '•-.'•           '              !"'•'•      •      -•   '
In general, organic Level M-l's emphasis is on reviewing positive (detected) :data.  The primary
question asked is whether or not a compound is present  If it is,  the next question addressed is
whether the compound is potentially from field or  laboratory-induced contamination.  If the
answer is negative, men the presence of the compound is considered confirmed, and the reported
concentrations are considered usable for some predefined data uses.  If information regarding
data quality and usability is required, then the emphasis is shifted accordingly to an evaluation
of data quality parameters, false negatives, and detection limits (Level M-2 oriM-1).  If legally
defensible data are necessary, then a full, CLP-equivalent data validation is performed (Level
M-3 or IM-2).  There is a definite focus at every step of the process. This allows for a clear
differentiation in the levels of data validation.

For guidance purposes, general data use categories and suggested levels of review are provided
in Table 1-3. However, it is important to note that the selected level of review will be specific
to the intended data use and specific project objectives.
                         Standard Operating Procedures

SOPs have been developed for Levels M-l, M-2, and IM-1 and are in the following sections
of this document:

      •     Volatiles (Appendix A)
      •     Semivolatiles (Appendix B)
      •     Pesticides/PCB (Appendix C)
      •  ,   Metals and Cyanide  (Appendix D)

-------
                             CONTENTS
                                                                  Page

Introduction

Glossary of Acronyms and Terms

Appendices (SOPs)

A.    Validation of Volatile Organic Analyte Data (Levels Ml and M2)

A-l. Manual Level Ml (for VOAs)
      1. Purpose and Applicability  . ..	... . . 1
      2. Quality Control Measures Checked	 ....	1
      3. Procedure ..-,..•*;;>*,.,.... •........ป-,-...-.;.	 .1
           3.1 Action Level Notification	 3
           3.2 Evaluation of Mass Spectra for Detected
                  Compounds		 .6
           3.3 Evaluation of Chromatograms . . .  . ;	 ... . ...... .  . 10
           3.4 Evaluation of Retention Times  ...................... 14
           3.5 Evaluation of Blanks  . .	 .  .	 16
           3.6 Sample Paperwork	17
                                Tables
Ml-VOA-1   QC Checklist for Level Ml CLP RAS Volatile Organics .-.-...	  .  2
Ml-VOA-2  Volatile Organic Analytes and Action Levels	  5
Ml-VOA-3   VGA Mass Spectral Evaluation Summary.	'..,.:..-	11
A-2.  Manual Level M2 (for VOAs)
      1. Purpose and Applicability	1
      2. Quality Control Measures Checked		3
      3. Procedure	 .  .	3
           3.1 Action Level Notification	7
           3.2 Technical Holding Times	10
           3.3 GC/MS Instrument Performance Check	12
           3.4 Initial Calibration  .	14
           3.5 Continuing Calibration	 18
           3.6 Blanks	 . .	.20
           3.7 System Monitoring Compounds (Surrogate
                 Spikes  . .	25
           3.8 Matrix Spike/Matrix Spike Duplicate	28
           3.9 Internal Standards	29

-------
CONTENTS (Continued)
                                                                  Page
           3.10 Reported CRQLs	 .		31
           3.11 Tentatively Identified Compounds		 32
                                Tables
M2-VOA-1  QC Checklist for Level M2 CLP RAS Volatile Organics .......	2
M2-VOA-2  VGA Qualifier Summary .. . . . ....	.... . . •> ;ป ., .'-i . . 5
M2-VOA-3  VGA Surrogate and Matrix Spike Quality Control Summary... . . .... 6
M2-VOA-4  Volatile Organic Analytes and Action Levels ....;.;. . ...w;v;.yj,. . . * 9
M2-VOA-5  Qualification of Volatile Analytes Based on technical Holding Times . .11
M2-VOA-6  Qualification of Volatile Analytes Based on System Monitoring
            Compound Recoveries.................  . . . . . :^.;.. , . .  .27


B.    Validation of Semivolatile Data (Levels Ml and M2)

B-l.  Manual Level Ml (for SVGAs)
      1. Purpose and Applicability  ........... . ..... ... . .  ....  ... . .1
      2. Quality Control Measures Checked	 ... :.., . .  . .,. . . 1
      3; Procedure V• • '• • •'.••••••,.."..........'.-........ •//;.',,;'.:..,... . .1
           3.1 Action Level Notification	 :............ ,3
           3.2 Evaluation of Mass Spectra for Detected
                  Compounds	  ^ .... .5
           3.3 Evaluation of Chromatograms  . .	^	9
           3.4 Evaluation of Retention Times  ...................... 15
           3.5 Evaluation of Blanks	 16
           3.6 Sample Paperwork	 . .  . •	17
                                Tables        v .
Ml-SVOA-1 QC Checklist for Level Ml	2
Ml-SVOA-2 Semivolatile Organic Analytes and Action Levels.	.4
Ml-SVOA-3 Mass Spectral Evaluation Summary	1Q
                : .              '     "                 ..'.''.

B-2.  Manual Level M2 (for SVGAs)                   ,
      1. Purpose arid Applicability	......'	 i
      2. Quality Control Measures Checked	3
      3. Procedures		3
           3.1 Action Level Notification  . .  .	9
           3.2 Technical Holding Times	11
           3.3' GC/MS Instrument Performance Check	13
           3.4 Initial Calibration		16
           3.5 Continuing Calibration	20
           3.6 Blanks	 22

-------
 CONTENTS (Continued)
                                                                    Page
            3.7  Surrogate Spikes	.  . .	28
            3.8  Matrix Spike/Matrix Spike Duplicate ...	........... 32
            3.9  Internal Standards ....:...............	 34
            3.10 Reported CRQLs		... .... .	, ..  . 36
            3.11 Tentatively Identified Compounds  ..... .;.;*ป ^.,.,.,.... . 37
                                 Tables
M2-SVOA-1  QQ  Checklist for Ijevel M2 .'. . .;. . ... . .;.:^&smasmwK8w:^.-.'.  2
M2-SVOA-2  SVOA^Qiiii^Smnm   .......... .Vv-^^sss^^j^^ ;... 5
M2-SVOA-3  SVGA Surrogate and Matrix Spike QC Summary '.-:,^:ws^'•<••'••:• :• •'•'• • • 7
M2-SVOA-4  Semivolatile Action Levels ... . . . . . . . .,;..'•^^^•^^m-i.•;-.. .   10
M2-SVOA-5  Qualification of Semivolatiles Based on Surrogate Recoveries  . .. . . . . 31
 C.    Validation of Pesticide/PCB Data (Levels Ml and M2)
'  '        .   '••'  •V;.;^f%^';.^' ' •/ ;•• .
 C-l. Manual Level Ml (forPesticides/PCBs)
      1. Purpose and Applicability  ..........	. '^^Sj^^fj^s....... I
      2. Quality Control Measures Checked ...... .:. .^^||fi^.. '. .'... . .  . 1
      3. Procedure  ..-..;.--:.;,:.,'.•,.......'....-...;. ^'ฅฃjฃ,'ฃ*.--..i'. .-.-'•. .  .1
            3.1 Action Level Notification	...... .,'i;^fv :..;.... ..3
            3.2 Evaluation of Retention Times', ... . ..i,.';-;'-'.^; >''..'..;	5
            3.3 Evaluation of Chromatograms ..... .-.i ., *. ; . .Cr% ... .  . . .  .7
            3.4 Evaluation of Blanks	 — ........\ .........  14
            3.5 Sample Paperwork	  19
                                 Tables
 Ml-PEST-1  Checklist for Level Ml PEST/PCB			.2
 Ml-PEST-2  PEST/PCB Action Levels.		.	4
 Ml-PEST-3  PEST/PCB Retention Time for Windows			8
 Ml-PEST-4  PEST/PCB Retention Time for Evaluation Summary	.9
C-2. Manual Level M2 (for Pesticides/PCBs)
      1.  Purpose and Applicability	 1
      2.  Quality Control Measures Checked	 . . .	3
      3.  Procedures	3
            3.1  Action Level Notification	  . 7
            3.2  Technical Holding Times	9
            3.3  GC/ECD Instrument Performance Check	11
            3.4  Initial Calibration	  16
            3.5  Continuing Calibration  .	 .	.18
            3.6  Blanks	.,	 .  21
            3.7  Surrogate Spikes	....;.	25

-------
CONTENTS (Continued)
            3.8 Matrix Spike/Matrix Spike Duplicate		27
            3.9 Pesticide Cleanup Checks		... . ... . . 28
            3.10 Reported CRQLs		,	 30
            •-.".-.  . /:.:.-';,.-. :•,  •   Tables ,   .        ''-.  ..'-.-
M2-PEST-1  QC Checklist for Level M2 .'. ............ .*..>,	2
M2-PEST-2  PEST$^ฃi^	:..... ....;,:.%.-> v;, ,.. .5
M2-PEST-3  ^PKTYPl^              '.,. '..... .••. .•^^Xjj^^xy*. • • •&
M2-PEST-4  PEST/piGB	.v.... ^.•:f:;..!.'.:J.. .8
      Attachment to Appendix C-2
               - ••.••'--"•.•^••X.-..-.'.V;..i-..
             ' •'" •' "•"""* *~''xttfy\/P.l*rJjฃpjฃ!>?--•&.*!?•*• ' • . '           ' ' '
D.    Validation of Inor^nlc Analyte Data (Level Ml)

      1.  Purpose^d^i^liclbiiity ,	;......... .1
       '   ''"'.''-,  ':-^ฃ:^ฃ'^:-l    '    '•' •    '•'  •    •'  .."''••• '.•>?.••''•.   . • '•  •  '•'••
      2.  QualityCoh^l/Measiireis.Checked ....'.-...	.... .>	2
       ' •  •'  2.1'':Krckieduie$-%^............ .•...... /....:.... ,>•:. '•.. '.•...... .2
            2.2 Reporting Requirements  ..................... . . :. . .4
            2.3 Action Level Notification  .... . . ...... r......... . . . . .6
            2.4 Holding Times ......	.	8
            2.5 Calibration .,........ . .... . .. .... ..... ..v^. ..... .9
            2.6 Blanks			.. T:	13
            2.7 ICP Interference Check Sample (ICS) ...... . .	14
            2.8 Laboratory Control Sample (LCS)	 16
            2.9 Duplicate Sample Analysis	 17
            2.10 Matrix Spike Sample Analysis	 . 18
            2.11 Furnace Atomic Absorption QC	 20
            2.12 ICP Serial Dilution	21
            2.13 Field Duplicates .			22
            2.14 Reporting Limit Verification	 23
            2.15 Sample Paperwork  ...	24
                                  Tables
IM-l-INORG-QC.    QC Checklist for Level Ml-CLP  RAS TAL Inorganics .	3
IM-1-INORG-AL     Target Analyte List Metals and Action Levels	7
IM-1-INORG-HT.    Holding Time Summary	(back)
IM-1-INORG-CAL.    Calibration Summary	J	 (back)
IM-1-INORG-SPK.    Summary of Spike Recovery /DuplicatePrecision	(back)
IM-1-INORG-QL.    Summary of Data Qualifiers	(back)
1M-1-INORG-IRDA   Inorganic Regional Data Assessment.	        (back)

-------
                 Glossary of Acronyms and Terms
Acronyms

%D •
%R
AA
AOB
APO
BPB
BNA
CADRE
CARD
CCB
CCS
ccy
CF
CLP
CLPRAS
CRDL
CRQL
CSF
CV
DAS
DFTPP

DPO
DQOs
DV
BCD
EICP
EMSL-LV
EPA
GC
GC/EC
GC/MS
GPC
HAZRAP
ICB
ICP
ICS
ICV
IDL
IRDA
IS
LCS
percent difference
percent recovery of spiked amounts of analytes
atomic absorption
analytical opei^Jions
                             ''^.vtt^s
Bromofluorobene(volatik instrument
(ปmputer aided dare^aew
Cl^P analytical .result^ database
cx>ntinumg
   •  .•'• k •*".  , -v'
calibration factr
contract
con
contract
computer required ;qitioiQ -Unoit
complete SDG file
Coefficient of Variation
delivery of analytical services .
Decafluorotriphenylphosphine (semivolatile instrument performance
  check)
deputy project officer .
data quality objectives
data validation
electron-capture detector .
extracted ion current profile
Environmental Monitoring Support; Laboratory—Las Vegas
United States Environmental Protection Agency
gas chromatography
gas chromatography/electron capture
gas chromatography/mass spectra
gel Permeation Chromatography
Hazardous Waste Remedial Actions Program
initial calibration blank
inductively coupled plasma
interface check sample
initial calibration verification
instrument detection limits,
Inorganic Regional Data Assessment
internal standards
library control sample

-------
MCL
ml
MS/MSD
MSA
m/z
NEESA
NET
QADS
  RDA
PEST/PCB
PRP
QA/QC
XJAPjP
OC
QL
QN
RAS
RIG
RI/FS
RPD
RPM
RRF
RRT
RSD
RT
SAS
SDG
SMC
SMO
SOP
SOW
SVGA
TAL
TCL
TIC
TPH
TPO
VGA
maximum contamination limit
milliliter
matrix spike/matrix spike duplicates
Method of Standard Addition.
the ratio of mass (m) to charge (z) of ions measured by GC/MS
Naval Energy and Environmental Support Activity
National Institute of Standards and Technology
organic analysis datasheet (Form.I)
                     Assessment
                     bibhehvls
                      rป-v-•.y . •
                        ility study
     .. .. Z^^^r--- ••• •
remedial project 'manager
     • v. . * - )••**• .'"V-l ' ' •••'. **
relative retention time
relative standard deviation
retention time     '
special analytical services
sample delivery group
system monitoring compound
sample management office
standard operating procedure
statement of work
semivolatile organic analyte
target analyte list
target compound list
tentatively identified compound
total petroleum hydrocarbons
technical project officer
volatile organic analytes

-------
                                   TERMS

Associated Samples        Any sample related to a particular QC analysis.
                          For example:
                          •     For ICV, all samples run under the same
                                calibration curve.
                          •     For duplicate RPD, all SDG samples digested/
                                distilled of the same matrix.

Calibration Curve          A plot of absorbance versus concentration of standards.

Case
                          collected over a given time period for a particular site.
                          A Case consists of one of more Sample Delivery
                          Group(s).

-------
               Appendix A
Validation of VolatUe Organic Analyte Data

     •  Subappendix A-l coven; Level Ml

     •  Subappendix A-2 covers Level M2

-------
                                                        Procedure no.:  Ml-VOA
                                                                    Revision:  1
                                                               Date:  06/30/1995
                                                                   • Page 1 of 18"
                                 Appendix A-l
             Validation of Volatile Organic Analyte Data
                             Manual Level Ml
                        1.  Purpose and Applicability

 This procedure provides step-by-step instructions to manually validate the volatile organic
 analyte '(VOA) data using the manual innovative data validation approach at Level Ml.
..The procedure is applicable tor^eJ'VOA-data^b
|fiojg;rain Statement of YYb^ (O3^^                                   CLP SOW
^specifications are essential in^ brd^|i^carr^out
.Data validated using this procedure are ^considered usable for the following types of
 purposes; however, the data users^jnustidecide .on  a case-by-case basis whether the
 procedure is suitable for their intended data uses. The suggested data uses are:

       •      Oversight of activities led by other parties

       •      Comparison to action levels

       •      Initial site investigation

       •      Contamination sources


                    2. Quality Control Measures Checked

 Table Ml-VOA-1 highlights the quality control (QC) indicators evaluated under this data
 validation procedure.


                                3.  Procedure

 The following subsections  describe for each of the QC indicators the acceptance criteria,
 location and retrieval of QC data, evaluation of the QC data, actions taken in the event the
 QC acceptance criteria are exceeded, and documentation of the  QC violations  in  a
 standardized report form.

-------
                                                         Procedure No.: Ml-VOA
                                                                     Revision:  1
                                                                Date:  06/30/1995
                                                                    Page 2 of 18
                              * Table Ml-VOA-1
                       QC CHECKLIST FOR LEVEL Ml
                       CLP RAS VOLATILE ORGANICS
             QC MEASURES

 InstrwnK
-------
                                                         Procedure no.: Ml-VOA
                                                                      Revision:  1
                                                                Date:  06/30/1995
                                                                     Page 3 of 18

Reporting requirements for Level Ml are:

       •     Hand annotate the Form I's, including
                   Data validation qualifiers
             -     Sample identification number
                   Sampling location

       •    " Provide a narrative that includes
             -     A statement mat defines the level of the data review, i.e., Ml
                   Major problems associated with analysis

       •     Include the following attachments
                   List of data validation qualifiers
             - -     Support documentation including forms that support assigning data
                  .qualifiers                                       :
                   Chain of custody form

3.1  Action Level Notification
             '           '*'•'.    '      •  •                  . '   •.
The purpose behind action level notification is to make the EPA Remedial Project Officer
(RPM) or the Site Project Officer  (SPO) aware of the potential human health risk at the
site.  In accordance with the Region m Hazardous Waste Division policy, the EPA RPM
or SPO must be promptly notified of any contaminant exceeding the established action level
or the 10-day health advisory limit. The data for contaminants exceeding the action levels
must be validated as a top priority and reported to the RPM or SPO.  Validation of the rest
of the data may then be completed normally.

3*1.1  Acceptance Criteria

EPA's Office  of  Solid Waste and Emergency Response has established 10-day  health
advisory limits or  the action levels for several organic compounds and elements of special
health  risk concern based  on  the Safe  Drinking Water  Act.   The  volatile organic
compounds and their 10-day health advisory limits are listed in Table Ml-VQA-2.   The
criteria for action level notification  are as follows:

       •     -The contaminant concentration must be equal to or above the established 10-
             day health advisory limits.

       •      Data for contaminants exceeding the action levels must be validated as a top
             priority.

       •      The  following  EPA personnel  must  be  notified  of the  action  level
             exceedances:

-------
                                                 Procedure no.:  Ml-VOA
                                                             Revision:  1
                                                        Date:  06/30/1995
                                                            Page 4 of 18


            EPARPMorSPO

      -     EPA Section Chiefs:   •

            -     Site Investigation (SI)
            "'  .' • ' •-  • '
            . —     Remedial

            — ,   Enforcement

            -     RCRA

      -•,'••' EPA Section Toxicologists:

            - •   ' Enforcement

            —     Superfund

                  RCRA

•     The remaining data validation should be completed per normal procedures.

•  '   Any  special  instructions  from the  Hazardous Waste Division  should be
      followed.                                        .

•     Records should be kept of the data review, action level notification and any
      follow-up instructions and actions.

-------
                                                         Procedure nor. Ml-VOA
                                                                      Revision:  1
                                                                Date:  06/30/1995
                                                                     Page 5 of 18
" Table Ml-VOA-2
VOLATILE ORGANIC ANALYTES AND ACTION LEVELS
Compound '.
Benzene

Chlorobenzene •.:
1 , 1-Dichloroethylene
trans-1,2- .^.
Dichloroethylene ;•':, J. '•', ..;..
l,24)ichtoroprppane"i;^ •?%
Methylethyl ketone
(MEK, 2-butanone) V ^
Tetrachloroethylene .
1,1,1-Trichloroethane .
Xylenes
Action
Level*
,'.Vv'-233-:. .'.'-.:
:.ฃpOO,V'v
-1,000
:^2;720,;;- '
3^90 ^v>
••••}&•**
-------
                                                          Procedure no.:  Ml-VOA
                                                                      Revision: 1
                                                                 Date:  06/30/1995
                                                                     Page 6 of 18
3.1.5  Reporting
Copies of Form I's can be used to highlight the contaminants above the action levels.  The
findings of the focused validation can be summarized in a memorandum,  and the  data
qualifiers resulting from focused validation may be written on  the Form I's,  The marked
up forms should be clarified that they, represent validation of only  the  contaminants
exceeding the action levels,and.not all data.

32 Evaluation of Mass Spectra For Detected Compounds;:
      •        ''!•*
The primary QC indicator checked in  Level Ml is  the  mass  spectra for the detected
compounds.  This indicator pertains to evaluating a compound's presence by matching its
•mass spectrum with a standard ;(knbwn) mass spectrum for the compound.'.No assessment
is made of the reported quantity of the compound or any quantitative  quality control
indicators that could lend some uncertainty to the reported value.

3.2.7  Acceptance Criteria _
                  "  : •   .*•- -:--  .'.'
The acceptance criteria ^for inass spectral matching  are  given hi the CLP Functional
Guidelines and are as follows:,
                . • :  .   •'     \
      •      All ions  present in the standard (known) spectrum at a relative intensity
             greater than 10 percent must be present in the sample mass Spectrum.

      •      The relative intensities of the qualifying ions (those above 10 percent relative
             intensity), must be within  +/-20 percent between  the standard and sample
             spectra, for example, an ion with a relative abundance of 50 percent in the
             standard  spectrum must be present between a relative abundance of 30 and
             70 percent hi the sample spectrum).

      •      Ions present in  the sample above 10 percent relative abundance must be
             accounted for.

3.2.2  Review Items

Form I's for each field sample, field blanks, and laboratory blanks  included in  a Sample
Delivery Group (SDG)  are necessary to compile a list of the detected compounds. Mass
spectra are then necessary for  each of the detected compounds  in a sample.  The required
mass spectra include both the sample spectra as well as the standard spectra.

With respect  to the blanks, it  may be beneficial to evaluate the blanks before the sample
mass spectra are evaluated.  If a compound is found to be a common contaminant, i.e., it
is present in any one of the blanks and also in the sample, the concentration in the sample
should be  evaluated with respect to  the highest blank contamination using the  5 (or 10)

-------
                                                           Procedure no.: Ml-VOA
                                                                        Revision:  1
                                                                  Date:  06/30/1995
                                                                       Page 7 of 18
               •
times criteria before proceeding with the elaborate mass spectral evaluations. The effort
required for the  sample  mass spectral  evaluation may  be significantly reduced  by
performing the blank evaluation first, The mass spectra for the blanks; however, should be
evaluated to ensure that the contaminants were identified properly.

All sample Form I's are generally located together in front of the data package. Copies of
the Form I's can .also be found in the detailed sample data package organized by sample;
i.e., the Form I and its substantiating raw data for a sample are placed together.  The raw
data include a quantitation report and mass spectra for each detected compounds in that
sample are.  A standard  (known)  mass spectrum, and  two sample mass spectra  (one
unaltered and another background-subtracted) are generally provided.  The presentation of
the mass spectra differs for different instrument manufacturers, but information necessary
for evaluating mass spectra is always provided in some form.  Hands-on experience with
different brands  of  mass spectrometers,  although  not essential, can  be helpful  in
interpreting the information.

Copies should be made of the  Form I's, preferably from the summary data package, if
included.- Otherwise, the forms can be pulled out from the raw sample data package.  All
detects should then "be highlighted with a marker or other convenient means.  The. raw data
should  then  be tagged for  every sample for evaluating the  mass  spectra  and the
chromatograms (reconstructed ion current profiles).

3.2.3  Evaluation of Mass  Spectra

Compare the highlighted Form  I's with  the tagged raw sample data.  Verify that sample
results are present for all of the samples listed on the chain-of-custody form.  Verify that
raw data are provided for every sample,  and a Form I exists for every sample represented
by raw data.  Verify on a sample-by-sample basis that mass spectra are provided for every
detected (highlighted) compound reported on  the Form I's.  A cross verification should
then be performed that all target compounds for which mass spectra are  provided  are also
reported as detected on Form  I's.            "

Mass spectral comparison routines based  on the evaluation  criteria  presented  earlier
(presence of major'ions and their relative intensities) are built into the instrument software;
therefore, performed automatically.  A listing of ions and their relative intensities is not
produced; rather a score based  on the extent of the fit of all criteria is produced.  The
score is based on a basis of 100 or 1000 for a perfect  match of the presence of major ions
and their relative abundances.  The score is printed on the quantitation report as a  "q"
value.  A score upward of 60 percent of the maximum (upward of 60 or 600) is generally
considered acceptable for positive identification for a compound.

Visual comparison of a mass spectrum involves looking for the base ion (the mass fragment
with the highest intensity), the parent ion (mass fragment equal to  the molecular weight of
the compound) and other characteristic ions representing removal of one or more functional

-------
                                                          Procedure no.: Ml-VOA
                                                                       Revision:  1
                                                                 Date:  06/30/1995
                                                                      Page 8 of 18

groups (such as -CH3, -CH2-, -Cl, -OH, or a combination thereof) from tfte parent ion, the
base ion or other ions. The removal of functional groups is often successive and produces
a fingerprint pattern for a type of compound; i.e., straight-chain hydrocarbons produce a
characteristic envelope of mass fragments separated by 14 mass units.  Similarly, aromatic
compounds produce a peculiar fingerprint. The base ion represents the most stable mass
fragment; therefore, it is always present. However, a parent ion often is unstable and may
not exist at all or may exist at a low abundance or relative intensity.              .

Compare  the standard (known) mass spectrum with the background-subtracted sample
spectrum  for the presence of the base, parent and  other characteristic  ions.  Although
background-subtracted and .unaltered mass spectra are generally provided, the; former type
of spectra are much cleaner looking due to the subtraction of column bleed or other broad-
based interferences; .therefore,  better  suited for comparison.  Generally, the presence of
major ions and overall matching of the fingerprint-pattern  between the standard and the
sample spectra can be considered satisfactory.  The  comparison is rather subjective, and
requires a trained-eye  to deduce the information.

Interference still may be present in a background-subtracted mass spectrum due to co-,
eluting compounds ( as compared to column bleed or broad-based interferences). Unless
the interfering compound is an isomer or an analog of the target compound in question, the
fingerprints produced by the target .compound and the interfering compounds can be quite
different.  Inseparable isdmers reflect a limitation of the gas chromaiography without any
recourse.  Such isomers are reported as ' 'total'', for example, total xylenes (combination of
ortho-, meta- and para-xylenes).  Analogous compounds generally have different retention
times; therefore, may  not be of much concern. Thus, co-eluting compounds do not pose
any problems in evaluating the mass spectra.

Seldom does a CLP  laboratory  incorrectly  interpret a mass spectrum,  but there  is  a
subjective element  to mass spectral interpretation.   While working within the  CLP
guidelines, the  laboratories  do have their own reporting  practices  at the limit of the
instrument's sensitivity.  Most of the  mass spectral identification problems occur near the
limit of detection where the differences in the relative intensities are not easily discernible.
Special attention should be paid in evaluating mass spectra for compounds detected at low
concentrations.   In addition, oxygenated compounds  such as acetone and 2-butanone
produce poor mass  spectra.  For such compounds, the  standard spectrum may show the
presence of the parent  ion, but the same may not be true of the sample spectrum.
Professional judgment should be exercised in evaluating such mass spectra.

Rearrangements and other side-reactions often occur inside a mass spectrometer.  These
phenomena produce mass fragments that are not easily accountable from the structure of
the parent compound.

-------
                                                          Procedure no.:  Ml-VOA
                                                                      Revision:  1
                                                                Date:  06/30/1995
                                                                     Page 9 of 18
3.2.4 Action
In the event the  mass spectrum of a detected compound does not at all resemble  the
standard spectrum or has extremely poor matching, the compound should be considered
undetected.  In this case the detected value should be changed to the CRDL for that sample
and  flagged  as undetected, "U."   (NOTE:   This action is in, contrast  to the EPA
Functional Guidelines which recommend rejection of data.  It is felt that  an outright
rejection of data is not justified.  If the mass spectrum produced at a target compound's
retention time does not match the standard spectrum, a conclusion can be drawn that the
target compound is not present and the mass spectrum may be due to something else.  In
such a case the target compound should be considered as undetected and data should not be
rejected.  Make sure  that the target compound in question does  produce a good  mass
spectrum by inspecting the standard mass spectrum from the calibration and the fit score on
the quantitation  report
-------
                                                         Procedure no.:  Ml-VOA
                                                                     Revision:  1
                                                                Date: 06/30/1995
                                                                    Page 10 of 18

       •     N—Tentative identification.  Partial proof of a compound's presence, but all
             identification criteria not met.

       •     U—No mass spectral match. Using professional judgment, the compound is
             considered undetected.

       •     R—No mass spectral match. Using professional judgment,  the compound
             data are rejectedfrom further use.

3.3 Evaluation of Chromatograms

The purpose behind evaluating the chromatograms is to get an idea regarding potential false
negatives, and gross analytical errors.  Checking for positive data as described under mass
spectral'evaluation does not offer any insight intodata that are not reported (i.e., reported
as nondetecis). Laboratory error <ซ• gross interference from other compounds could be the
reasons for erroneous reporting.  The gas chromatograms or the reconstructed ion current
(RIG) profiles are the primary tools' used for the false negative evaluation under Level ML

3.3.1  Acceptance Criteria

There are no EPA-established criteria for evaluating chromatograms for false negatives.
The criteria used for evaluating chromatograms are based on good laboratory and scientific
practices, and these are not hard and fast requirements.  The suggested evaluation criteria
are:'      .      .      ••'''.-.

       •      There should not be any significant peaks in the chromatograms that are not
             accounted for  as  TCLs or TICs.   Significant  peaks are  those with  a
             minimum peak height of 10 percent of the closest internal standard.

       •      The chromatograms  should ideally  have base-line  resolution  between
             adjacent peaks.  Also, there should not be broad (unresolved) envelopes in
             the chromatograms.

       •      There should not be abrupt shifts in the baseline.

       •      There should not be peak tailing or sharp rise in the peak fronts.

-------
             Case No.:
             SDG No.:]
      Data Reviewer:	
          Site:
Procedure No.:  Ml-VOA
           Revision:  1
      Date: 06/30/1995
         Page 11 of 18
              Table Ml-VOA-3
VGA MASS SPECTRAL EVALUATION SUMMARY
Dates Analyzed:
Instrument ID:
^sr .
OwpoouL ,
CaotomeOiMat

1

T • • • ....
A . • ' '•' '"•• •"

a '• • . '•. . '••••-:
7 ". •' ' .•.-•- '•' ."•
8 ' ' • ' : ' '" ''*''•'
Q. . :•'••' :'' '-. • -V •:_''••
10 - • • .'.. '•' •'•
t
V)
Biwnbme&MB -
Vinyl Chloride
(*)
Chlotoefl^ne
Memyfcne CUoride
Acetone
Carbon Dbalfide
1.1-Dichloroethene
1,1-Dichloroethane
(*)
(#)
1,2-Dichloroethene (total)
Chloroform
(*)
1 ,2-Dichloroethane
2-Butanone
1,1, 1-Trichloroethane
Carbon Tetrachloride
Bromodichloromethane
1 ,2-Dichloropropane (*)
cis-1 ,3-Dichloropropene
Trichloroethene
Dibromochloromethane
1 , 1 ,2-Trichloroelhane
Benzene
AR
trans-1 ,3-Dichloropropene
Bromoform
(#)
4-MethyI-2-Pentanone
2-Hexanone
Tetrachloroethene
1,1,2,2-Tetrachioroethane (#)
Toluene
*)(AR)
Chlorobcnzene (<0(AR)
Ethylbenzene
Styrene
Xylenes (total)
ซ)(AR)
(AR)
(AR)

i •
•>. • .. _ _ • ......
ซ=• •-• 	 - - •' .:•:••••

<...••••. ' ' ' • •.-••.-•• . .• . • .
i; • -.-• . . •• : • :• . .. -•'••:-..' ".•'vsv.T-ifte; '•• -y •••••;•• ••••-.
7-.' -. .'• • ' - :•••: •• ' • .-/'.' • " '•• *> •'. '-. : .
'ซ' ' ••.•'•• -• • • = ' .••••.-,•.:••'•.
,0-.7-::'---A'-.; • .•'••• - •,. • •' ---•••'••••:••=, .'.:•.' •./•••;"..••.:
^io/'- : •••- '•". .•••..• .•-.-•;.:,-•.,>, :Vv. ' .'.•• : •••
'; '• ' ; .- ^.''•'••^"^••'•'•M^'j •••"'; Sample Identifier :. ^"•-' v-!vv vv';''.:::^''-' . - ' '
^ 1
MS






;


























RT

































•v "->•••
MS
..•v
V. ."



• •'-
•


























RT
'.i ••
'.-,•'••••
.,.






























*^^
MS
>i'?>
•';:-fc:;






• •
























RT
•V.;;?
''".- '•































'^-4^
MS






-'


























RT
:;v':..'
-- -



•.



























'•-*•• ~
MS

































'RT
'_;
>































6
MS

































Rt










'






















'•'.i •'.
MS



•r





























RT
"•;*.





1


























,'/ซviy..:'-
MS
/?- rv-
•.'-•'•'ซ

•• ''!i'
. •/:


. ~
.-
























RT
•K-;
-V.vv


•. ป•.


^:,^
"—
























.>:-9- '•
MS
••,-V';V
\-v'

-?• -




•'.
























RT

•/-.'

. • '





























10
MS


:. *
• -






















.






RT


































-------
                                                          Procedure no.: Ml-VOA
                                                                       Revision:  1
                                                                 Date:  06/30/1995
                                                                     Page  12 of 18
3.3.2  Review Items
Chromatograms, or the RIC profiles for each sample are necessary for the evaluation.  The
RIC profiles can be found  in the front of the raw data package for each  sample.   The
quantitation report for each sample is also necessary to retrieve either the retention times or
information regarding scan numbers to compare with the peaks on the Chromatograms.

3.3.3 Evaluation of Chromatograms
Visually inspect the chrontegrMjs for. all peaks that appear to be at least 10 percent in
height of the nearest internal standard. Compare that the scan number or the retention time
that appears on the x-axis of the.RIC profile with that listed on the qiiantitation report for
the TCLs or the Ubrary search record for the TICs.  Make sure that all significant pealcs
are accounted for.

Also observe the RIC profile for peak resolution between adjacent peaks:  Poor peak-to-
peak resolution is indicative of degrading performance of the gas chromatographic column.
The values obtamed from a degrading system are prone to be inaccurate.   Obviously
certain isomeric or homologous compounds are difficult to separate.  But generally, there
should beat least 90 percent valley between the neighboring peaks.

Inspect the RIC profile for broad, -unresolved envelopes. These are generally indicative of
outside interference  from a series  of homologous compounds  such as- straight-chain
hydrocarbons.  Especially, assess the interference with the internal standards and surrogates
under the envelope using the expected area counts.  The standard values that are far from
the expected values may be indicative of potential problems with the TCL detection or
quantification.
                                           '
Inspect the RIC profile for abrupt shifts in the baseline.  Such  shifts are  indicative of
problems with instrument sensitivity or  leakage in  the system.  The area counts obtained
from shifted baseline are inaccurate, or  even the detection of a TCL at low concentration
may be missed.
                   I        "••...
Rapid peak rising or peak tailing indicate problems with the gas chromatographic column,
such as depleted stationary phase on  the column, decomposition of the stationary phase or
creation of active sites. Again,  a visual inspection of the RIC profile will yield information
on the shape of the peak.

3.3.4 Action

Professional discretion must be used when evaluating and qualifying data based on the
chromatographic evaluations. An experienced chemist can generally infer the magnitude
and the frequency of; the problem from the RIC profile.  If the problem appears to  be

-------
                                                          Procedure no.: Ml-VOA
                                                                      Revision:  1
                                                                Date:  06/30/1995
                                                                    Page 13 of 18

systematic,  then  it  should  be  brought to the  laboratories  attention  and resolved.
Intermittent problems may or may not require any action.  The following guidelines are
suggested when acting on RIC profile observations:

       •      Any unaccounted TCL peak with area equivalent to or greater than the
             lowest reportable limit for the sample must be brought to the laboratory's
             attention and resolved. Any unaccounted non-TCl^
-------
                                                         Procedure no.:  Ml-VOA
                                                                     Revision:  1
                                                                Date:  06/30/1995
                                                                -   Page 14 of 18

             problem. For example, the magnitude of a drop in tne oaseline below the
             zero line may not be estimated and could be very significant. On the other
             hand, a drop  that yields a baseline still  above zero can be  put in a
             perspective with the original baseline and a general appearance of the entire
             RIC profile.

       •     The .problems with peak symmetry are indicative of system degradation, and
             should b^bjtpught: to the attention of the laboratory for a corrective action.
             Professional judgment should be used when and if qualifying any data due to
                                 -,. First the problem should  be defined in terms of
             perastenoe||bipuhout the chromatogram and also from sample to sample.
                             ^shapes and area counts for  the internal and  surrogate
                               i evaluated to see if the problem could have affected
                                ; and/or quantification.  Data qualification  may, be
                                     area counts are acceptable.

3.3.5 Reporting

The Mass Spectral Bva^uation Form (Table Ml-VOA-3) presented earlier may be used to
note any brief comments bii tiie'chromatographic evaluations. The comments may be noted
against each sample identifier^  For more descriptive comments, a separate sheet may be
used.  -  .    •.-•••-• •^'•••'-' •••-••••••-                   '     ;  ;   -

3.4 Evaluation  of Retention Times

While not unequivocal in identifying a compound, the retention times are quite helpful in
confirming the  presence of a  detected compound.   Matching  of mass spectrum  and
retention time of sample with those of a standard yields higher credibility and confidence
level to the detection in the sample.  On the other hand,  not matching the retention time
may or may not invalidate the  detection. If a mass spectral match is made beyond any
doubt, poor matching of the retention time may not have adverse impact on the detection.
If mass spectral matching is unacceptable or only partial, and the retention times do not
match, then a strong doubt can be cast on a compound's presence.
                  ii                             '                         •
3.4.1  Acceptance Criteria

The criteria for retention times are specified in the EPA's  functional guidelines as follows:

      The relative retention times (RRTs) must be within +/-0.06 RRT units  of
      the applicable internal standard RRT.

-------
                                                         Procedure no.: Ml-VOA
                                                                      Revision:  1
                                                                Date:  06/30/1995
                                                                    Page 15 of 18
3.4.2 Review Items
Quantitation reports  for the sample and  continuing calibration  are required  for the
evaluation. These are located in the raw data for samples and standards.' A copy of the
continuing calibration quantitation reports may be made or the reports pulled out from the
raw data to facilitate a comparison with the sample quantitation reports.  -

3.4.3 Evaluation Procedures

For the detected compounds, determine the relative retention time for the compounds by
dividing their retention times with the retention time of their associated infernal standard in
the samples as well as in the applicable continuing (or initial) calibrations.The sample
RRTs must fall in range of standard RRT +/- 0.06. units.

3.4.4 Action

Action for retention time evaluation requires professional discretion. Action taken must be
based on other data such as  mass spectra and not on retention times alone.  .The following
actions are suggested for several potential situations.        :

       •      Acceptable matching  of the mass  spectra  and the  RRTs—No action
             suggested.

       •      Acceptable matching of the mass spectra but poor matching of the  RRTs—
             No action suggested,  but  the  cause of  retention  time shift  may be
             investigated.  Check to see if similar discrepancies are observed in  other
             parts of the chromatogram.  Often high concentration of a compound will
             cause shifts in the retention times for other compounds in the nearby eluting
             region, but shifts in retention times in other areas may not be observed.
             Shifts all through the chromatogram may be indicative of an erratic system,
             such as  flow  rate fluctuations, poor temperature regulation, restriction or
             leakage in the system.  There are likely chances that the sample values in
             such conditions may  be inaccurate.   Positive data may be qualified as
             estimated, "J," if deemed essential.

3.4.5  Reporting

The Ml-VOA-MS Form (Table Ml-VOA-3) presented earlier may be used to record any
problems in the RRT matching. For the detected compounds in each sample, the calculated
RRTs may be recorded under the "RT" column. A data qualifier code may  be added to
the values exceeding 0.06  RRT and requiring qualification, such as  "0.15J," for a
compound with a difference of 0.15 RRT  units and where  a professional judgment to
estimate the data is exercised.

-------
                                                          Procedure no.: Ml-VOA
                                                                       Revision:  1
                                                                 Date:  06/30/1995
                                                                     Page 16 of 18
3.5 Evaluation of Blanks
Laboratory blanks and field blanks have a profound impact on false positives reported in
samples; i.e., compounds reported as positive detects but not originating from the samples
themselves.  Cross contamination from the sampling equipment, incidental contamination
from the field conditions or contamination from the laboratory  equipment or general
environmental are likely sources of false positives in the samples.  .
          •'        ..:'••.'.:--!- ซ•"'•'•'-.'.•",    '  ' : "           .       . .   '• '    ' :'.     •."
3.5.1 Acceptance  Criteria! >
                 ./;^-;^^;-'>. ..   '.-        ''.'•'.-  ;  :  ••''>•  -.' '
Criteria for blank evaluation are specified in the EPA's functional guidelines.   In addition,
Region m has some  additional requirements  modifying the guidance.  The acceptance
criteria for blanks apply equally to any type of blanks associated with either sampling or
analysis, such as trip blanks^ rinsate blanks, field-or' bottle blanks, and laboratory method
blanks.   While' there are several criteria for evaluating  the blanks, the .only criterion
applicable to Level Ml is the comparison of the blank and sample concentrations.  This
criterion  is:   '• '   '•  .-•;.^v.;^^.y/,.;..•  '             ••..'•."   •'•  '.•••'.
        •  ,  '   '•' •  •^•;'^.-'-HryVffi'-;;.'',' '-.  .                   •:•.••  --. '  . '
      For  common  contaminants,  such  as  methylene  chloride, acetone  and
      2-butanone, the sample concentration must be  minimally 10 times the blank
      .concentration to be considered a positive detect.  Other contaminants must
      be present in the sample at or above 5 times the blank concentration.   The
      blank with the  highest concentration must be used (this is, if the laboratory
      blank has the highest concentration of methylene chloride, and the equipment
      rinsate blank has the highest concentration of acetone, both of these blanks
      must be used to qualify the respective contaminants).  All blanks should be
      checked and the highest concentration of contaminants in any blank should
      be used for data qualification.

3.5.2 Review Items

Data requirements and data retrieval procedures for blanks are the same as  those for the
field samples because the blanks as well as field samples are validated similarly. Form I's,
mass  spectra,  chromatograms, quantitation reports, etc., are essential  for performing a
validation of the blanks first.                          .

3.5.3 Evaluation  Procedure

Validate  the blanks the same as the field samples,   detailed validation procedures are
described above  under  appropriate  sections.    Use  the  validated blank results  for  a
comparison  with  the  sample results.   Make certain  that the samples and blanks are
evaluated on the same basis of sample weight or volume, dilution factors, moisture content,
etc.   Use the 5 (or 10)  times the highest blank concentrations for qualifying the  sample
data.

-------
                                                          Procedure no.:  Ml-VOA
                                                                      Revision:  1
                                                                 Date: 06/30/1995
                                                                     Page 17 of 18
3.5.4 Action
                                                                  -i-.
                                                                 meeting the blank
If the sample  concentrations  do not meet  the  criteria of 5  (or  10) times the blank
concentration, the sample results should be considered essentially undetected (or as not
detected  substantially above the levels reported in the blanks); therefore, flagged "B" in
accordance with the Region ffl data validation guidelines.         _

3.5.5 Reporting

Form I's may be used to write the "B" data qualifier for the.data not
'•criteria.       '  •'  .    '•;•'•  ••(•''  • .  -    '.    ' .      ./.  ••   ' •'••••

3.6 Sample Paperwork  "  ;      '       ;

The purpose for evaluating the sample paperwork is to determine that {he samples being
validated are indeed the ones taken from the site, and have not been tampered with.
Accurate sample identity is of paramount  importance in substantiating "the sample data.
Without unequivocal sample identity and chain-of-custody procedures, the sample data may
not be defensible or enforceable.                             •'••:"
Under the current CLP contracts, the original paperwork (i.e., the purge package or the
administrative record) is included in the data package from the laboratory. It is assumed
that the data validator is not privy to the original paperwork;  therefore^ the evaluation
criteria and  procedures described below apply only to the documents that are generally
included in the data validation package.  These documents are the chain-of-custody forms
and Region HI Shipping Record.

3.6.1  Acceptance Criteria

Criteria for acceptability or authenticity of the sampling paperwork,  document control and
chain-of-custody have been established by the National Enforcement Investigations Center*
(NEIC), in support  of the CLP.  Overall criteria are too numerous and subjective to be
discussed here, but the criteria that apply to data validation are as follows:

       •      The chain-of-custody  form should  be properly and  completely  filled  out
             including the sampler signatures, date and time of sampling, sampling station
             identification, analyses requested, traffic numbers, tag numbers, etc.  These
             data are minimally  required to confirm the authenticity of the sample and its
             data.

       •      The chain-of-custody must be maintained at all times.  The custody transfers
             between different parties must be signed and. dated.;

-------
                                                          Procedure no.:  Ml-VOA
                                                                      Revision:  1
                                                                 Pate:  06/30/1995
                                                                    Page 18 of 18
3.6.2 Review Items
A copy of the chain-of-custody form is,essential to confirm the identity of the samples. In
addition, the Region m Shipping Record is essential to identify the field QC samples.  The
chain-of-custody and Shipping Record are generally located in front of the data package.

3.6.3 Evaluation Procedure
                   '.  •.••.••.•v-J.s-F:1/'?:-'-1-"' .-..-•,                          ••  .
   •  '  •'     -•   •. ' • '-'#^$^;$&^;  ' •...• •    •     ••       ' •'     •••'"'••:''•.
Ensure that the chain-of-custody fprm was signed and dated by the samplers, and a time
and date were entei^/for>sainpleVq)liection. The laboratory copy of the chain-of-custody
must have the signature of the laboratory sample custodian.  Any errors on the form should
have been  crossed out with a single line through the entry.  Verify  that all collected
samples  have unique  station identification^ traffic numbers  and sample  tag numbers.
Ensure that the Region HI Shipping Record correctly reflects the information on the chain-
                                   "   ''
3.6.4 Action

The action to be taken in qualifying the data is highly dependent on the nature of the
problem.  Some errors in paperwork are practically unavoidable in real situations.  An
effort should be made to reconcile the differences by cross checking the field notebooks
against the sampling paperwork. Occasionally, the samplers may forget to sign the chain-
of-custody;  however,  the field  notebooks  may  amply describe  the  sampling event.
Problems are also inevitable in noting or cross-referencing sample tag numbers and traffic
numbers.  Generally, there are several alternate sources of information to substantiate or.
refute the problem.

3.6.5 Reporting

Any discrepancies found in the paperwork must be immediately brought to the attention of
the EPA RPM or SPO.  Clearly define the problems in  a memorandum to the responsible
parties.  Attach  marked copies of the chain-of-custody forms to substantiate the findings.

-------
                                                         Procedure No.:  M2-VOA
                                                                      Revision:  1
                                                                 Date:  06/30/1995
                                                                     Page 1 of 34
                                 Appendix A-2
             Validation of Volatile Organic Analyte Data
                              Manual Level M2


                         1. Purpose and Applicability
                       :.':-.V^i-'!:'i:'.->..y j^-.     ' " • •.  . ••• '-•..••.-'•••••;'•'••'•-"<ฃ:
This procedure provides step-by-step instructions to manually validate the volatile organic
analyte (VOA) data using the /manual innovative data vah'dation approach at Level M2.
This approach focuses on the use of information contained on the GLPforms andla review
of chromatograms as summarized in Table M2-VOA-1. The procedures are based on
modifications to Regions Hi's National Functional Guidelines for Organic Data Review.
.'       .     •      *^ •    .'-"-. .--. •     •_ -•'.. .        f         •   • .'   *^ .   .""-.:.  •  .   . .
The procedure is applicable -to rlhie VOA data obtained  using '
Program Statement of Work (CLP SOW).  Hard copy data conforming to the CXP SOW
specifications are essential in order, tp cany out the procedure.^/  V. c  ;l!:-:; j"  /    :   ;

Data validated using  this procedure are  considered usable  for the following types of
purposes; however, the  data -users must  decide on  a case-by-case basis  whether the
procedure is suitable for their intended data uses. The suggested data uses are:

       •     Oversight of activities led by other parties

       •     Comparison to action levels

       •     Initial site investigation

       •     Contamination sources

       •     Nature and  extent of contamination

       •     Preliminary risk assessment

       •     Risk assessment with known high levels of toxics

       •     Feasibility study

       •     Preliminary design

             Treatability study                             US' EPฃ,aH.ead3uarters Library
                      33                                   Mail code 3201
                   i         T  *•                      120ฐ Pennsylvania Avenue NW
                   cleanup verification                       Washington DC 20460

-------
                                                      Procedure No.:  M2-VOA
                                                              Revision No.:  1
                                                             Date: 06/30/1995
                                                                  Page 2 of 34
                              Table M2-VOA-1
                     QC CHECKLIST FOR LEVEL M2
                     CLP RAS VOLATILE ORGANICS
Action Level NotificaticaiW
        Tune
Initial Calibration (RRF)
•• '  •  •  '•
Initial Calibration (%RSD)
       ''  ''' •'
  .    .     ...
Continuing Calibration C^Rl^
      •.      ..   •.. .•
Continuing Calibration (%D);
Labbratoiy .Blank
MS/MSD (%R,RPD)
Internal Standard Area
Field Blank
Sample Paperwork
Holding Time
Retention Time
Surrogate Recovery
Dilution Factor
Moisture Content
Mass Spectra
Chromatograms
Raw Data
'm^~:
  X1"5 •fc
  ... '':
                                      X
                                              M3
                  '  X
                    X
                                                            Comhinfid
".X^tte
•' ..'.j.;-S-'.-
                             X
                                             C3
                                           "...-'X
                                           . , " -

-------
                                                        Procedure No.:  M2-VOA
                                                                    Revision: 1
                                                               Date: 06/30/1995
                                                                    Page 3 of 34
                   2.  Quality Control Measures Checked

Table M2-VOA-1 highlights the quality control (QC) indicators evaluated under this data
validation procedure.                   .  .
                               l!3.  Procedure
                          *&**&%ฃ'':, :--\  •
The following subsections describe for each of the QC indicators the acceptance criteria,
            W .     ...._, '•ฃ '. -" ..'>iwi*.-i:;--~ - -       '     .           . - -        *. • -           *
data requirements and retrieyaTpTXJC data, evaluation of the QC data, actions taken in the
event the QC acceptanceIcnte^iureexceeded, and documentation of the QC violations in
a standardized report
The volatile data i^i^mentsltq^;.(ฑe6ked are listed below:
                   ,  - ' "_,"_ '• • - ,V*"- f'.^.v";.' >**'• - '    '   - •
       3.1    Action Level Notir|cation

       3.2    Technical Holding Times (CCS - Contractual holding times only)
                           .\  " ..
       3.3    GC/MS Instrument Performance Check (CCS)

       3.4    Initial CalibrationXCCS)

       3.5    Continuing Calibration (CCS)

       3.6    Blanks

       3.7    System Monitoring Compounds (CCS)

       3.8    Matrix Spike/Matrix Spike Duplicate

       3.9    Internal Standards

       3.10   Reported Contract Required Quantitation Limits (CRQLs)

       3.11   Tentatively Identified Compounds

Two forms have  been  developed  to  assist in  the performance  and documentation of
implementing  Level M2.  The first form,  M2-VOA-QUAL, summarizes  holding time,
calibrations, blanks, surrogates, and internal standards.  The second form, M2-VOA-SPK,
summarizes surrogate and matrix spike quality control checks. These forms are presented
in Table M2-VOA-2 and Table M2-VOA-3.

-------
                                                           Procedure No-  M2-VOA
                                                                        Revision:  1
                                                                  Date:  06/30/1995
                                                                       Page 4 of 34
Reporting requirements for Level M2 are:

       •     Hand annotate the Form I's, including
                   1 Data validation qualifiers
                    Sample identification number
                    Sampling location
             .    •                    ..
             Provide a i narrative ithat includes
                    A statement that defines the level of the data review, ie. M2
                                    problems assockted with the analysis,
                    Highlight issues that may have affected detection limits
                    ,' ' . .:-V-r:':-Vi;i"''v.5!v'^-' .V.' ' '  •'    •                 '    .•••-,.
                    1 -..•.'.':{ .>ป;'&. .•u,*-a&*tฃ  , .                  .      '      "    ..-•
             •  •   .-.,-' -.,1v:iJisr--.^j!-.-.iซ.":-  .           .  .            •.  ...  ••-•-.':•.•.
             Include the following attachments  ^ *• • .         .         '"""
             -   A   Ustof^^Walidation qualifiers                      :
                  • •' ,.  ••* '•"-'• '••j.'.J'ป^*W''*'1-.
-------
                                        Case No..
                                        SDG No.:.
                            Data Reviewer::	
                                      Site:	,
                        Procedure No,:  M2-VOA
                                       Revision:  1
                                 Date:  06/30/1995
                                      Page 5 of 34
           Table MB-VOA-2. VOA Quali&r Summary (CaEbntipos, Blanks, Haldmt lime, Surrogates, Internal Standards)

  Dite(i) Analyzed:
  Samples Analyzed Within 12 Houn of
  Tune?   	Vea   	No
  Tune CMC? _Yes  _No



  Method Blank ID: .
Sample Identifier:
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
Hold Time
Out, days
AR










All










Stindinfi
s
SI










tlTPOffftt
S3.

••








*M
S3

•








: ซ.>1
ISTDt
IS1










B2


' •






•
IS3










                               Date:
Time:
Compound: SPCC (I)
CCC(ป)
Aromatic (AR)
Chtomfmtfiiine ' (f\
BrramiK*ane
Vinvl Chloride f*>
ChlorMlhape
Methvlene Chloride
Acetone
Caibon Dinilfide
1.1-Dichloraethene f*t
1.1-Dichloroethane (fl
1.2-DichloroetheneYtotan
Chloroform <*>
1 2-Dichloroethane
2-Butflnone
1 . 1 . 1 -Trichloroethane
Carbon Tetrachloride
Bromodichloromethane
1 .2-Dichlorooronane (*\
cis-1 ,3-Dichloroorooene
Tnchloraelhene
Dibromochloromethane
1 . 1 .2-Trichloroethane
Benzene AR
trans- 1 .3-Dichlorooronene
Bromoform (If)
4-Methvl-2-Pentanone
2-Hexanone
Tetniihloroethene
1.1.2.2-Tetnchloroethane (ft
Toluene f*VA!H
Chlorobenzene (f)(AR)
Ethvlbenzene (*)(AR)
Stvrene fARI
Xvlenes ftotaH (AK\
Initial Gd.
RRF<.05

































*RSD>30

































Coatin^C*
RRF<.05

































ซD>25

































Blanks
Mett. Trio

































,






- .,

























OoaBfiers
<+/•)
; •


































>i
s
f
i
i
s
i
2
I
S
1
3
TICs Reported In BUnk(i):
                              Reported as:
      El
                                                                           fog/kg. ug/Lt

-------
              Case No.:_
              SDG No.:.
    Data Reviewer:	
             Site:	
Procedure No.:  M2-VOA
              Revision:  1
        Date:  06/30/1995
             Page 6 of 34
Table MJ-VOAJ. VOA Surrogate and Matrix Spike Quality Control Summary
Sample Identifier
(Acceptance Ranee, %R):
1.
2.
3.
4.
5. .. ' " '
6.
7.
8.
9.
10.
11.
12.
13.
Aoueoui Sample Reeoveriet, %R
SI
88-110






--• .,
' t



1 - ,,

52
86-115






.





•" •
S3 '
76-114




(



1
•



Soil Sample Recovery. %R
SI
84-138










\


S2
59-113













S3
70-121













Qualifiers ซ•/-)

•

. •-











             SI • Tohwne-dS, S2 •Bromofluorobeozeae,S3 • l^-diehlofoe1faane-d4

                  MS/MSD Data Sdmmarv
SPCCd)
CCC(ซ)
Spike ComponndAromatk (AH)
Metric SiAe. ttttmirv %R

lUnge
Actual
Matrix Spike Dupfcate,
RecoTerv. %R
Range
Actual
MS/MSD Precision, RPD
Range
Actual
Qualifiers
<+/•)
AQUEOUS SAMPLES
1,1-Dichloroethene (*)
Trichloroethene
Benzene AR
Toluene (*)(AR)
Chlorobenzene (WAR)
61-145
71-120
76-127
. 76-125
75-130





61-145
7M20
76-127
76-125
75-130





14
14
11
13
13










SOIL SAMPLES
1,1-DicWoroethene (*)
Trichloroethene
Benzene AR
Toluene (*)(AR)
Chlorobenzene (D(AR)
59-172
62-137
66-142
59-139
60-133





59-172
62-137
66-142
59-139
60-133





22
24
21
21
21


'








-------
                                                         Procedure No.: M2-VOA
                                                                      Revision:  1
                                                                Date:  06/30/1995
                                                                     Page 7 of 34
Codes Related  to  Quantitation (Can be  used  for  both  positive results and  sample
quantitation limits):

       J      =     Analyte present. Reported value may not be accurate or precise.

       K     =     Analyte present Reported value may be biased high. Actual value
       L     =     Analyte present. Reported value may be biasiedtow^Actual value is
                    expected to be higher.  •             -", v^'v7%S:;:; V;

       UJ    =     Not detected, quantitation limit may be inaccurate or imprecise.
                             *ฃf-*'         '•*'•'•''•     ;'.   ,-:f^fi$^
       UL   =     Not detected/ quantitation limit is probably higher.
               .  ••   •.        " -<^'^
Other Codes

       Q     =     No analytical result
  •   "  r             '-.  ,.
       *     =     Results reported from diluted analysis.

3.1  Action Level Notification

The purpose behind action level notification is to make the EPA Remedial Project Officer
(RPM) or the Site Project Officer (SPO) aware of the potential human health risk at the
site.  In accordance with the Region m Hazardous .Waste Division policy, the EPA RPM
or SPO must be promptly notified of any contaminant exceeding the established action level
or the 10-day health advisory limit.  The data for contaminants exceeding the action levels
must be validated as a top priority and reported  to the RPM or SPO. Validation of the rest
of the data may then be completed normally. •

3.1.1  Acceptance Criteria

EPA's Office of  Solid Waste and Emergency Response has established  10-day health
advisory limits or  the action levels for several organic compounds and elements of special
health  risk concern based on the  Safe  Drinking  Water Act.   The  volatile  organic
compounds and their 10-day health advisory limits are listed  in Table M2-VOA-4.  The
criteria for action level notification are as follows:

       •     The contaminant concentration must be equal to or above the established 10-
             day health advisory limits.

-------
                                            Procedure No.:  M2-VOA
                                                         Revision:  1
                                                    Date: 06/30/1995
                                                         Page 8 of 34
Data for contaminants exceeding the action levels must be validated as a top
priority.
The following EPA  personnel  must  be  notified  of  the action level

          .
            ^^investigation (SI)
          •••'  --.i'''-' .'•    •''
      EPA Section Toxicologists:
 ^    •':'•'"';' V;--^;;|^M;V\ '.'. "'
      r-   V Enforcement

      —     Superfund

      -      RCRA

The remaining data validation should be completed per normal procedures.
               ' •      "                                  '
Any special instructions from the Hazardous Waste Division should be
followed.

Records should be kept of the data review, action level notification and any
follow up instructions and actions.

-------
                                                         Procedure NOM M2-VOA
                                                                      Revision:  1
                                                                Date:  06/30/1995
                                                                     Page 9 of 34
Table M2-VOA-4
VOLATILE ORGANIC ANALYTES AND ACTION LEVELS
Compound "..,". ...-.
Benzene • ^ '
Chlorobenzene ; .'
1 , 1-Dichloroethylene
trans-1,2- . * • ,
Dichloroethylene
1,2-Dichloropropane ' .
Methylethyl ketone ' .
(MEK, 2-butanone)
Tetrachloroethylene
1,1,1-Trichloroethane
Xylenes
Action
"gl*vel* v .
%^t;233. ';••'
%4l;&0q' • •
1,000
v*'! . ป. ป ' " •
, 2,720
' - / • i
90
,^7,500;: v
f-.-f-f ',. . .;
'34,000
35,1)00
7,800
. ' • :CompounbV"^j|ฃ^
Carbon tetrachloride l^^

1,2-Dichloroethane ^ 4v c-l
cis-l,2-Dichloroethylene
Dichloromethane>^>3^
• f ^ i -ti ;-j A^-"?.^f^
(in^thylene chloride) ^ni
• Ethylbenzene ;^^^
'Styrene. • '.x,yV^^|^
-"-'•• v • ' • — •<'-v'ti*-i^i~/'
Toluene ' ^-W^ff?
Vinyl chloride
— ' '.. "•"'-
Action
iatcyd* •
K&o.-:
fl^wo:-;-'
|S-i,ooo .
!W**.--.r.— ;•-. ' •
sfSjS'?.* ซAA-~'
• ;
tfei'ioo'- ; '
^-5ft nnn
$*^?yw
W-^A-.: • •'. •
f-Pe^oo" ; •
i 2,600
-w.- .—
*A11 units are ug/1.
3.1.2 Review Items

All data required to perform the complete Level M2 validation, as detailed in the following
sections are necessary for carrying out action level notification.  The location of the data
and their retrieval procedures are also discussed below.

3.1.3 Evaluation  Procedure

The evaluation process preceding  action  level notification will primarily  consist of
comparing the results on Form I's with  the action levels presented in Table M2-VOA-4.
Following the identification of the contaminants exceeding the action  levels, focused data
validation should  be  performed  using  the criteria,  and procedures described  in the
appropriate sections below.

3.1.4 Action
The action resulting from focused data validation will be the notification of action level
exceedance to the personnel identified above in Section 3.1.1.

-------
                                                        Procedure No.:  M2-VOA
                                                                     Revision:  1
                                                                Date:  06/30/1995
                                                                   Page 10 of 34
3.L5 Reporting

Copies of Form I's can be used to highlight the contaminants above the action levels.  The
findings of the focused validation can be summarized in a memorandum, and the data
qualifiers resulting from fojcn^^dation may be written on  the Form I'S.  The marked
up forms should be clarlfie^l^that- they  represent  validation of only  the contaminants
 *       •        .-•.-. v.- . ;:iiซ 'A- -Til *•*<:ป:-/>•. •'•••.*       •        .     • -••,.
exceeding the action

3.2  Technical Efo
  .
The objective is tt>
from time of coUection tb^
              •
,„,.  .-.    '
3.2.1 Acceptance Crtie
    -  •.•    f  •• ... v:,;^fe
                                    of results based on the holding time of the sample
                                  ysis.  -   '
                                      •'
        .                ._, _______ w .. .
Technical requireme^vfi^^u^lpholding times have only been established for water
matrices.  The holding time^^fOT^oils (and other non-aqueous matrices such as sediments,
oily wastes, and sludge) are^omn^tly^under investigation. In Region iti, a 14 day holding
time will be applied to all nim2aqueous  samples.  When soil holding time criteria are
established and. available, jfhe/j^rocedure'for qualifying soil samples will be re-evaluated.

The holding time criteria for water samples, as stated in the current 40 €FR Part 136
(Clean Water Act) is as follows:                                 .
      For non-aromatic  volatile compounds in cooled (@ 4ฐC) water samples, the
      maximum holding time is 14 days from sample collection.

      Maximum holding times for purgeable aromatic hydrocarbons in cooled (@ 4ฐC +.
      2ฐC), acid-preserved (pH 2 or below) water samples are 14 days from sample
      collection.                          •           .
                   'i . -          •
      Water samples that have not been maintained at 4ฐC (ฑ 2ฐC) and/or preserved to a
      pH of 2 or below should be analyzed within 7 days from sample collection.  If
      insufficient ice is used to ship samples, the laboratory may receive samples with no
      ice left in the cooler.  Under these circumstances, the temperature of the samples
      may exceed 4*C.

It is further required that volatile compounds in properly preserved non-aqueous samples be
analyzed within 14 days of sample collection for all volatile compounds.

3.2.2  Review Items
Form I VOA, EPA Sample shipping log and/or chain-of-custody.

-------
                                                         Procedure No.:  M2-VOA
                                                                     Revision:  1
                                                                Date: 06/30/1995
                                                                    Page 11 of 34
3.2.3 Evaluation Procedure

       Technical holding times are established by comparing the sampling dates on
       the EPA Sample Traffic Report with dates of analysis on Form I VGA
       Examine the sample records to determine if samples were preserved.' If it is
       indicated that there were problems with the samples, then the ihtegiity of the
         ..  . •  •   ;•'..•'.-^'.t^rU -ir^&akc -!• '.' ••.  •, '-: ..•   •  ""•. -;. ., --fcj|.^iJift*feSfeS*ฅ''ia' ซ-•'
       sample may have^be^^mpromised and  professional judgmelKShould be
       used to evalu^^e^effec^of the problem on flie sample
3.2.4 Action
       If technical holding times are exceeded, qualify the sample tesiili
       see Table M2-VOA-5).5- ;           V         : "i:"^

             If there is no evidence that the samples were
             technical holdmg times exceeded 7 days,
             and sample quantitation limits with "UL" for    ; s
             professional judgment to  determine  if and how''*
             compounds should also be qualified.           Jp
  ws.  (Also
ฅ&?• •
     and the
   with "L"
        Use
     volatile
             If the  samples  were properly preserved but the technical  holding times
             exceeded 14 days, qualify all positive results with 'lATiand  all sample
             quantitation limits with "UL."
Table M2-VOA-5
QUALIFICATION OF VOLATILE ANALYTES BASED ON TECHNICAL
HOLDING TIMES
Matrix
Water
Non-aqueous
Preserved
No
Yes
No/Yes
> 7 Days
All Aromatics*
None
None
14 Days
All Compounds
All Compounds
All Compounds
'"Reviewer should use professional judgment to determine if data for additional
compounds require qualification.
2.    If technical holding times are grossly exceeded (e.g., by greater than two times the
      required time for  volatiles) either on  the first analysis or upon re-analysis, the
      reviewer must use professional judgment to determine the reliability of the data and
      the effects of additional storage on the sample results.   Should the reviewer
      determine that qualification is  necessary, non-detected  volatile target compounds

-------
                                                        Procedure No.:  M2-VOA
                                                                    Revision:  1
                                                               Date:  06/30/1995
                                                                  Page 12 of 34
       may be qualified unusable (R).  Positive results are considered bias low and are
       qualified with "L".

3.     When there are other quality control problems in conjunction with' exceeded, holding
      . times (such as suspected;laboratory contamination), the reviewer should follow the
       hierarchy ofqualifim^^ particular, if for any reason me reviewer doubts the
       presence of a cpmpo^                  should display only tiie^'/B" or "R"
       qiialifiet and i^^                      is because no netjdirejc^cmx)f bias can be
       inferred  imder these^cbnditions.  When results are reported by the laboratory as
       below the CR^^thip^Lig qualifier is used over the "J" qualifier.

4.     Customized;^                      should be filled  to show holding time
       performance (oneTp^i|^H6ur period per instrument).
3.3  GC/MS InstrumentrPerformance Check
Gas chromatograpn/rnass spectrome^^         instrument performance checks (formerly.
referred to as tunmg) are performed to ensure mass resolution, identification, and to some"
      ••  .    •    .. ^" ft. .+},$, f -'-trJi5'-',v •'$ฃ' •<.,-.'•••-         •   • •        •  r ••;•",•• t-v •    •
degree,  sensitivity. VTheป^tem:iire not "sample specific,   Qmformance is determined
using standard materials, mereibre, these criteria should be met in all circumstances.
            -•  '.';',:•;':.  ; ^ 'K-.! -'.         ''•.'.:'        •  . •'•' •
3.3.1  Acceptance Criteria                                    :~-

The  analysis of the instrument performance check solution must be performed at the
beginning of each 12-hour period during which samples or standards are analyzed.  The
instrument performance check, bromofluorobenzene (BFB) for volatile analysis, must meet
the ion abundance criteria given below.

      Bromofluorobenzene (BFB)
                 !•
      m/z         Ion Abundance Criteria

       50          8.0 - 40.0% of m/z 95
       75          30.0 - 66.0% of m/z 95
       95          Base peak, 100% relative abundance
       96          5.0 - 9.0% of m/z 95
      173          Less than 2.0% of m/z 174
      174          50.0- 120.0% of m/z 95
      175          4.0 - 9.0% of mass 174
      176          93.0 - 101.0% of m/z 174
      177          5.0 - 9.0% of m/z 176

-------
                                                           Procedure No.: M2-VOA
                                                                        Revision:  1
                                                                  Date:  06/30/1995
                                                                      Page 13 of 34
NOTE:       All ion abundances must be normalized to m/z 95, the nominal base peak,
              even though the ion abundance of m/z 174 may be up to 120 percent that of
              m/z 95.

3.3.2 Review Items

Form V VGA.
                •••'•"'". •*?'> " :>i

3.3.3 Evaluation[tin^dK !^

1.     Compare the data presented  for each Instrument Performance Check (Form V
       VGA) with each mass listing submitted to ensure the following: ''*|g;&;;;.':;_
             Form V VOA is present and completed  for each  12-hour period during
              V. . ••^>Y""ft'atf*'-'-' , ' 'j   '     '   -••-.'•  '• •••• .r.*-V*ti*-fซ^rV^.--.:.  •    ฐ'
             which samples were analyzed.           '           \:^:v:5y.::^,rv
                   •'..•"•.-:.'• ••'•A,^.'-'-;    •.       •:'.'•  • -:.'';• •>?i*"i.1>KfS.J5'**'''??-'

             The appropriate number of significant figures has teenreported (number of
             significant, figures given for each ion in the ion abundance criteria column)
             and that rounding is correct.  (See SOW for requirements).  1;.
                                            r
2.     Verify that the mass assignment is correct and that the mass listing is normalized to
       m/z 95.                                      .              
-------
                                                         Procedure No.:  M2-VOA
                                                                      Revision:  1
                                                                 Date:  06/30/1995
                                                                     Page 14 of 34
3.     If mass assignment is in error (such as m/z 96 is indicated as the base peak rather
       than m/z 95), classify all associated data as unusable (R).

4.     If ion abundance criteria are not met, professional judgment may be applied to
       determine  to  what extent the data may be utilized.   Guidelines to  aid  in  the
       application, of professional judgment to this topic are discussed as follows:

             The most ^important factors to consider are the empikcal|j^ulte ; that .are
                •' .  ;••••••••;• v *' •- . •'<• •• •   •               "  •  .  -  ' '•-^•^ซ*^i<-^^^V>'?&t *•-•ป••
             relatively insensitive to location on the chromatographic|prqffl.e^and the type
             of instrumentation.  Therefore, the critical ion abimdaince^cn^a for BFB
             are the m/?; 95^(5, 174/175, 174/176, and 17jS/177|^^5f|lie relative
             abundances,of $t]JQ. and 75 are of lower importanoej:fe:^^fe^;:       -

5.     If the reviewer has reason to believe that instalment performance check criteria
                 . •. .  ..I-:',...:,.- . i, . ,.:";'.•-. •           • •       •••*:'.- ";4ป-frtf>'t.'-\"-\ ^  ,-
                 . . . ...  . , . ,..                          .       .-     ,
       were achieved using techniques other than those described m n.p.4^then additional
       information on the/instrument performance checks shouW • b^^obtained.   If the
       techniques employed are found to be at variance with the contract requirements, the
       performance and procedures of the laboratory may merit ey^^qaX^Qmcerns or
       questions regarding laboratory performance should be noted for TPO action in the
       ORDAS form.   For example, if the reviewer has reason ^to believe that an
       inappropriate technique was used to  obtain background  subtraction (such as
       background subtracting from the solvent front  or from anbther^region of the
       chromatogram rather than the BFB peak), then this should be noted for TPO action
       in the ORDAS form.

3.4  Initial Calibration

Compliance requirements for satisfactory instrument calibration are established to ensure
that the instrument is capable of producing acceptable qualitative and quantitative data for
compounds on  the volatile target compound list (TCL). Initial calibration demonstrates that
the instrument  is capable of acceptable performance in the beginning of the analytical run
and of producing a linear calibration curve.
                                          r            ' ,
3.4.1  Acceptance Criteria
                   '<:                              '           •
1.     Initial calibration standards containing both  volatile target .compounds and system
       monitoring compounds are analyzed at concentrations of 10, 20, 50, 100, and 200
       ug/L at the beginning of each analytical sequence or as necessary if the continuing
       calibration acceptance  criteria are not  met.  The initial  calibration  (and any
       associated samples and blanks) must be analyzed within  12 hours of the associated
       instrument performance check.

-------
                                                         Procedure No.: M2-VOA
                                                                      Revision:  1
                                                                 Date:  06/30/1995
                                                                    Page 15 of 34
2.     Separate initial calibrations must be performed for water samples (or medium level
       soil samples) and for low level soil samples. The calibration for water samples and

       medium level soil samples is performed with an uhheated purge and the calibration
       for low level soil samples is performed with a heated purge;        :  /
     '.-   '  '   : ••.-• ^^^^jy'. • '     .'.-''   '   .'" '.   '•  .';'•:.-•• ;7S''V •"'.'
3.     Initial c^bration);standani Relative Response Factors (RRFs)  for volatile target
       compounds and:^^m€^utoring compounds (surrogates) must ,be greater than or
       equal to 0.05. ((^ntra^tual initial calibration RRF criteria are listed in Appendix
       A\       •  \'^$&$*&3&'^--      •  .      '         .-.-•• . ••  •' •^•,: •'•  '•  '
      .-*•/••    "   .   , .v'AV^Ci;''-'-;?.•"•.' 'Y •           '         •    .-...._ .,.":.•; •  •

4.     The Percent Relative Standard  Deviation (%RSD) from the initial calibration must
       be less than or equal to 30.0% for all compounds. (Contractual calibration %RSD
       -ซป'••  •  ซ._."4ivi-H'ii''vfev,.j.-':.'...      '    ' '. '       •  . ., .•'-..". r-:.;;jv  '  • •.
       cntena are listed m Appendix A).
                    '-   ~'
3.4.2 Re}

Form VI VGA and chromatbgrams.
3.4.3 Evaluation Procedures

1.     Verify  that  the  correct concentration of  standards were  used; for the  initial
       calibration (Le., 10, 20, 50,  100, and 200 ug/L for water).       ^

2.     Verify that the correct initial calibration was used for water and medium level soil
       samples (i.e., unheated purge) and for low level soil samples (i.e., heated purge).

3.     If any sample results were calculated using an initial calibration, verify  that the
       correct standard (i.e., the 50 ug/L standard) was used for calculating sample results
       and that the samples were analyzed within 12  hours of the associated instrument
       performance check.

4.     Evaluate the initial calibration RRFs and RRFs for all volatile target compounds and
       system monitoring compounds (surrogates):

             Verify  that for all  volatile  target  compounds  and  system  monitoring
             compounds, the initial calibration RRFs are greater than or equal to 0.05.

NOTE:      Because historical, performance data  indicate poor response and/or erratic
             behavior, the volatile compounds listed below have no contractual maximum
             %RSD criteria. Contractually they must meet a minimum RRF criterion of
             0.01; however, for data review purposes,  the "greater than or  equal to
             0.05" criterion is applied to all volatile compounds.

-------
                                                        Procedure No.:  M2-VQA
                                                                     Revision:  1
                                                                Date:  06/30/1995
                                                                   Page 16 of 34
             Volatile Target Compounds Exhibiting Poor Response

                   •      Acetone
                   5      2-Butanone
                          Carbon disulfide
                          Methylene chloride
                          MM
                                      .
             NOTE:  Compounds in bold are system monitoring compounds.
S.    Evaluate the %RSD for all  volatile target compounds and  system monitoring
      compounds:  v;  >     V    .    .       :  '     '   '. '.'. '"." '•^'•'.'•.:''  •
                   !: 'v       -\  .  '".              '      ' '  . ' '    ''-"'•'
      a.     Verity that all volatile target compounds have a %RSD of less than or equal
             to  30.0%.  The contractual  criteria for an acceptable initial calibration
             specifies that  up to any 2 volatile target compounds  may- fail to  meet
             minimum RRF or maximum  %RSD as  long as they have RRFs that are
             greater than or equal to 0.010, and %RSD of less than or equal to 40.0%.
             For data review purposes, however, all compounds must be considered for
             qualification when the %RSD  exceeds the +. 30.0% criterion.
                                    /
      b.     If  the %RSD is greater  than  30.0%, then the reviewer  should  use
             professional judgment to determine the need to check the points on the curve
             for the cause of the non-linearity.  This is checked by eliminating either the
             high point or the low point and recalculating the %RSD.

3.4.4  Action
                   •                                      /           '
1.    All volatile target compounds,  including the   "poor performers"  listed above
      (excluding the system monitoring compounds) will be qualified using the following
      criteria:

      a.     If the %RSD is greater than 30.0% and all initial calibration RRFs greater
             than or equal to 0.05, qualify positive results with "J". Non-detects are not
             qualified.   When the %RSD is grossly exceeded (i.e., > 50%) use
             professional judgment for qualifying non-detects as "UJ".

-------
                                                          Procedure No.: M2-VOA
                                                                        Revision:  1
                                                                  Date:  06/30/1995
                                                                      Page 17 of 34


       b.     If any initial calibration RRF is less than 0.05, qualify positive results that
             have acceptable mass  spectral identification with  "L", and  non-detected
             analytes as unusable, "R".

2.  At the reviewer's discretion, a more in-depth review to minimize the qualification of
    data can be accomplished by considering the following:   ^••••'-^y^??-.

    a.       If any of the required volatile compounds have a %RSD greater than 30.0%,
             and  if elimlnaiirig dther the high or the low pomt-bf the curve does not
             restore the %RSD to less than or equal to 30.0%:      ^ ^

             i.  :    Quaiify;|X)sitive results for that opmpound(s) with ;**J".  '

            • ii.     No qualifiers are needed for volatile target compounds that were not
                   •detected. J  If  the  %RSD is  grossly exceeded  (U.,  >50%),
               .-'-.•    professional judgment is used to qualify non-detects with *'UJ".
                                      '                                '
    b.       If the Mgh point of the curve is outside of the linearity ^criteria (e.g., due to
         .    saturation):^ **%<-% •   '••-.'•   •   .'_',..  ,."""•'; "-"•''•   •
                          '\ ' •                       , '
             i.      No qualifiers are required for positive results in the linear portion of
                    the curve.     '                              • '.\--~-

             ii.     Qualify positive results outside of the linear portion of the curve with
                    a "J".

             iii.     No qualifiers are needed for volatile target compounds that were not
                    detected.    If  the  %RSD  is grossly  exceeded  (i.e.,  >50%,
                    professional judgment is used to qualify non-detects with "UJ".

    c.        If the low end of the curve is outside of the linearity criteria:

             i.      No qualifiers are required for positive results in the linear portion of
                    the curve.

             ii.     Qualify low level  positive results  in the area of non-linearity with
                    tซTปป
                     J '                         X

             iii.     No qualifiers are needed for volatile target compounds that were not
                    detected.    If  the  %RSD  is  grossly  exceeded  (i.e.,  >50%),
                    professional judgment is used to qualify non-detects with "UJ".

-------
                                                          Procedure No.:  M2-VOA
                                                                      Revision:  1
                                                                 Date: 06/30/1995
                                                                     Page 18 of 34


    NOTE:  If a, b, or c options are used, a description of the process must be clearly
             stated in the data review narrative.

3.  If the laboratory has failed to provide adequate calibration information, the designated
    representative should contact the laboratory and request the necessary information.  If
    the information is not available, the reviewer must use professional judgment to assess
    the data.     v;
                                      .                 ..           .,.,    .
4.  The potent^ effl^ion^the data due to unacceptable calibration criteria should be
    noted in the data review jnarrative.                            /

5.. If calibration criteria are exceeded, this should be noted on the ORDAS.
  • •'•    .  _; • '    •  . v'iV&'ivi.W,.*'.": •  ' '••
3.5  Continuing Calibration
    •  •-••"._  !'...'-'-#^l?^-vV•':••'' '
Compliance  requirement} for ^satisfactory instrument calibration are established to ensure
that the instrument is capable of producing acceptable qualitative and  quantitative data.
Continuing calibration^establishes the 12-hour relative response factors on which the
quantitations are based and checks satisfactory performance of the instrument on a day-to-
day basis.       ;    "': ••••-.'

3.5.1  Acceptance  Criteria.

1.  Continuing  calibration  standards containing both, target  compounds and  system
    monitoring compounds are analyzed at the beginning of each 12-hour analysis period
    following the analysis of the instrument performance check and prior to the analysis of
    the  method blank and samples.  The continuing calibration may either be a part of the
    initial calibration or run independently on another 12-hour analysis period. .
                  ป'    • •                i                  .     •
2.  The continuing calibration RRF for volatile target compounds and system monitoring
    compounds must be greater than or equal to 0.05.             .       ••

3.  The percent difference (%D) between  the initial calibration RRF and the continuing
    calibration RRF must be within ฑ 25.0%.

3.5.2  Review Items

Review Items:  Form Vn VOA and chromatograms.
                  ii

3.5.3  Evaluation Procedure

1.  Verify that the continuing calibration was run at the required frequency and  that the
    continuing calibration was compared to the correct initial calibration.

-------
                                                         Procedure No.:  M2-VOA
                                                                     Revision:  1
                                                                Date:  06/30/1995
                                                                    Page 19 of 34


2.  Evaluate the continuing calibration RRF for all volatile target compounds and system
    monitoring compounds:

             Verify that all volatile compounds and system monitoring compounds meet
             the RRF specifications.
NOTE:       Because nistpnc^ iperformance data indicate;:poor\r^ipoijse;and/or erratic
             behavior,vthe compounds listed in Section 3.4.3, above, have no contractual
             niaximum^D "criteria.  (^ntractiiaUy mey must meet a mmimum  RRF
            . criterion bl^pipl^hbwever, for data review purposes^the' "greater than or
             equal to 0.0^" criterion is applied to. all volatile compounds.      .

3.  Evaluate the^ %D;;J>e^^a|matial caUbratioiuRRF and continuing calibration RRF for
    one or more r-—-^ii-'a/-v**:* '
                                within ฑ 25.0% for aU volatiletorget compounds and
             system monitoring compounds.  Note those compounds which have a  %D
             outside thejat 2^.0% criterion.  The contractual OTteria for an acceptable
             contmumg^calibration specifies that up to any 2 volatile target compounds
             may foil to^ineet minimum RRF or  maximum %D as long as they have
             RRFs that are greater than or equal to 0.010, and %D of less than or equal
             to 40.0%.  For data review purposes, however, all compounds must be
             considered for qualification when the %D exceeds 4he ฑ 2570% criterion.

3.5.4  Action

1.  The reviewer should use professional judgment to determine if it is necessary to
    qualify the data for any volatile target compound.  If qualification of data is required,
    it should be performed using the following guidelines:

    a.        If the %D is outside the ฑ 25.0% criterion and the continuing  calibration
             RRF is greater than or equal to 0.05, qualify positive results  with "J".
    b.        If the %D is outside the ฑ, 25.0% criterion and the continuing calibration
             RRF is  greater than  or equal to  0.05, no qualification of  non-detected
             volatile target compounds is  necessary.   If the %D  is grossly  exceeded
             (>50%), professional judgment may be used to qualify non-detects with
             "UJ".

    c.        If the continuing calibration RRF is less than 0.05, qualify positive results
             that  have acceptable  mass  spectral  identifications  with  "L" or  use
             professional judgment and include justification for not qualifying the data in
             the data review narrative.

-------
                                                          Procedure No.:  M2-VOA
                                                                       Revision:  1
                                                                  Date: 06/30/1995
                                                                      Page 20 of 34
    d.        If the continuing calibration RRF is less than 0.05, qualify non-detected
              volatile target compounds as unusable, "R".

2.  If the laboratory has failed to provide adequate calibration information, the designated
    representative should 'contact the laboratory and request the necessary information. If
    the information is jfot available, the reviewer must use professional judgment to assess
    the data. -:-
3.  The potential -effectsrion^the data due to unacceptable calibration criteria should be
                 • '. ;  '.'•••f'- '••• ซ-". ';. •"'•' .                             .•',.''".'•
    noted in the data review narrative.                               ,v
      '                           '        '                      '           '
4.  If calibration criteria fareexceeded, this should be noted in the ORDAS.

3.6.  Blanks
                  . i. i^"Yv>k<
                  . • .-.,vJC".->.'
                 '•:\??i$$f^*$/'$?&:•''     ' •'                  '    '••'."••'.'•'•'
The purpose of -lalloio^^^^^cliS) blank analysis  is to determine the existence and
magnitude of oohtaiqiinM^^esulting from laboratory (or field) activities. The criteria for.
evaluation of blanks apply,to;any blank associated with the samples (e.g., methods blanks,
instrument blanks,; trip blanks,rand equipment blanks).  If problems with any blank exist,
all associated data must be carefully evaluated to  determine whether or not there is an
inherent variability in the data, or if the problem is an isolated occurrence not affecting
other data.

3.6.1   Acceptance Criteria

1.  No contaminants should be found in the blanks.

2.  A  method blank analysis must be performed after the calibration standards and once
    for every 12-hour time period beginning with the injection of BFB.

3.  The method blank must be analyzed on each GC/MS system used to analyze samples
    for each type of analysis, i.e.,  unheated  purge (water and medium level soil) and
    heated purge (low level soil).

4.  An instrument blank should be analyzed after any  sample that has exceeded the initial
    calibration  range  for  any  given  compound  to  check that  the blank  is  free  of
    interference and the system is not contaminated.

3.6.2   Review Items

Form I VOA, Form IV VOA and chromatograms.

-------
                                                          Procedure No.:  M2-VOA
                                                                       Revision:  1
                                                                  Date: 06/30/1995
                                                                      Page 21 of 34
3.6.3  Evaluation Procedures
 1.   Review the results of all associated blanks on the forms and chromatograms to evaluate
     the presence of target and non-target compounds in the blanks.

 2.   Verify that a method blank analysis has been reported per matrix, per concentration
     level for each 12-hour time period  on each GC/MS system used  to analyze volatile
     samples.   The reviewer can use the Method Blank Summary (FormTV VGA) to
     identify the samples •associated with  each method blank.       •  "  :;    :

 3.   Verify that the instrument blank analysis has been performed following any sample
     analysis where a target analyte(s) is  reported at high concentration^).:

 3.6.4 Action  ;       :   ;::/v
            ; .  .  •  '•^••^.' •."•.ฃ;,:" tiซV7'*\ :     •  •'•        '•         ".':-.';
           '  -   ..  v.--V:^'-fe;^;tfv  '                 '..-', ., ,';,:-i.- .  •
 If the appropriate blanks were Jiot analyzed with the frequency described in Criteria 2, 3,
 and  4,  then me data reviewer should  use professional judgment  to determine if th&
 associated sample data should be qualified. The reviewer may heed to obtain additional
 information from the laboratory.   The situation should be noted for TPO action on the
 ORDAS  form.         ;                                            !

 Action regarding unsuitable blank results depends  on the circumstances .and origin of the
 blank.  Positive sample results should be reported and qualified "B", if the concentration
 of the compound in the sample is less than or equal to  10 times (lOx)  the amount in any
 blank for the common volatile laboratory  contaminants (methylene chloride, acetone, and 2-
 butanone), or 5 times (Sx) the amount for other volatile target compounds.  In situations
 where more than one blank is associated with a given sample, qualification should be based
 upon a comparison with the blank having the highest concentration of a contaminant.  The
 results must not be corrected by subtracting any blank value.

 For qualification purposes, consider all blanks  in a case associated with all samples.

 Field blanks measure contamination introduced not only  in  the field but also from  the
 laboratory.  In  general, evaluation  of the impact on specific sample results is .handled the
 same as  with  laboratory  blanks.   The reviewer should  use caution   in  attributing
 contamination  to the  field as opposed  to laboratory  sources.   However, when field-
 introduced contamination is suspected, it  is  helpful for the reviewer to consult the sampling
 group to  identify possible sources and prevent future reoccurrences. Verified field sources
of contamination should be noted in the data  review narrative.  If a field  blank has  the
 highest concentration of a contaminant, then all samples in the associated case are qualified
 "B", using the 5x and  lOx  rule.   Other field blanks  associated with the case are not
qualified.

-------
                                                         Procedure No.: M2-VOA
                                                                      Revision:  1
                                                                Date:  06/30/1995
                                                                    Page 22 of 34
Specific actions are as follows:
1.  If a volatile compound is found in a blank but not found in the sample, no action is
    taken.  If the contaminants found are volatile target compounds (or interfering non-
    target compounds) at significant concentrations above the CRQL, men this should be
    noted for TPO action in the ORDAS.
         ....                      .                             i _
2.  Any volatile compound detected in the sample (other than the common volatile
    laboratory O)ntaminants)/4hat was also detected in any associated blank, is qualified
    4 'B",  when  the sample  concentration  is  less  than five  times (5x)  the  blank
    concentration.  For: common volatile laboratory contaminants,  the results are qualified
    "B",  when  the sample  concentration  is  less  than 10 times  (lOx)  the  blank
    concentration.-'   .:^:^/soV;:. "•• "      '• •   ^ .• •       •  •  '    ••••'./"''  .  •   -

3.  The reviewer should note that blanks may not involve the same weights, volumes, or
    dilution factors  as the associated samples.   These  factors must be taken into
    consideration when applying the "5x" and "lOx" criteria, such that a comparison of
    the total amoimt of contamination is actually made.

    Additionally, there may be instances where little or no contamination was present in
    me associated blanks/ but  qualification of the sample is deemed necessary.  If the
    reviewer determines that the contamination is from a  source other than the sample,
    he/she should qualify the data. Contamination introduced through dilution water is one
    example.  Although it is not always possible  to determine, instances of this occurring
    can be detected  when contaminants  are found in the diluted sample result but are
    absent in the undiluted sample result. Since. both results are not routinely reported, it
   ' may be impossible to verify this source of contamination.

4.  If gross contamination exists (i.e., saturated peaks by GC/MS), all affected compounds
    in the associated samples should be qualified as unusable (R) due to interference.  This
    should be  noted  for TPO action in the ORDAS if the contamination is suspected of
    having an effect on the sample results.

5.  If inordinate  numbers  of  other target compounds are found at low  levels in the
    blank(s), it may be indicative of a problem and should be noted for TPO action in the
    ORDAS form.

6.  The same consideration given to the target compounds should also  be given  to
    Tentatively Identified Compounds  (TICs), which are found in both  the sample and
    associated  blank(s).  (See VOA Section XHI for TIC guidance.)

-------
                                                           Procedure No.: M2-VOA
                                                                        Revision:  1
                                                                  Date:  06/30/1995
                                                                      Page 23 of 34
7.  If an instrument blank was not analyzed following a sample analysis which contained
    an analyte(s) at high concentration^), sample analysis results after the high
    concentration sample must be evaluated for carryover.  Professional judgment should
    be used to determine if instrument cross-contamination  has affected any  positive
    compound identification^).  If instrument cross-contamination is suggested, then this
    should be noted for TPO action if the  cross-contamination is suspected of having an
    effect on the sample results/Sample results which are pos^^a|tifects of carry-over
    should be flagged as unusable (R).
8.  When there is convincing evidence that contamination is  restricted to a particular
    instrument,  matrix, or concentration level, the Sx/lOx rule will :only be applied to
    compare contaminated blanks to certain associated samples T(as <|iposed to all samples
    in the case).  Some examples are as follows^  -•       •'•'             '
                    '        ''"''            '
    Column bleed (siloxanes) may be localized to a particular instrument
           '                     '   '     '             '
                                  .
    Methanbl extractions in the medium soil volatile analysis ^jrotocol can give rise to,
    contaminants that are not seen in the low-level aqueous analyses. 'y

    Common laboratory  contaminants, such as methylene  chloride, are generally  too
    unpredictable to safely assume contamination is restricted to a particular instrument,
    matrix, or concentration leveU                               -x '•-.--

9.  For benzene and/or  toluene,  the reviewer may identify that the observed laboratory
    contamination is attributable to a specific, regular, and predictable process (such as
    trap bleed), which results in a constant 1 or 2 ppb instrument level concentration in all
    runs (both samples and blanks).  In this situation, the reviewer may want to consider
    and flag certain results as tentatively identified,  "N", as opposed to  "B", if the
    sample instrument level is clearly greater  than the consistent level of contamination
    detected in blanks and other samples. (This particular situation supersedes the Sx/lOx
    rule.)                                ,

10. The following are examples of applying the blank qualification guidelines.  Certain
    circumstances may warrant deviations from these guidelines.  Any deviations  must be
    clearly stated in the data review narrative.

-------
                                                         Procedure No.:  M2-VOA
                                                                     Revision: 1
                                                                Date: 06/30/1995
                                                                    Page 24 of 34
Example 1:   Sample  result  is greater  than  the Contract Required Quantitation Limit
             (CRQL), but is less than the 5x or lOx multiple of the blank result.
                                                             Rule
                      Blank Result
                      Sample Result                   6Q ?    v    30
                      :Final Sample Result              60B         30B
             In the example for the "lOx" rule, sample results less than 70 (or 10 x 7)
             would be qualified "B".  In the case/of the "5x" rule, sample results less
             than 35r(dr5 x;7)fwquld be qualified "B".       .

Example 2:   Sample result is less than the CRQL, and is also less than the 5x or lOx
             multiple of the blank result.                   .

                                                             Rule    .
                      Blank Result                     6         .6
                      CRQL                           5           :5
                      Sample Result                    41          4J
                      Final Sample Result               4B          4B

                   Note that data are  reported as 4B, indicating that the qualitative
                   presence is not confirmed.

Example 3:   Sample result is greater than the 5x or lOx multiple of the blank result.

                                                             Rule
                                                      iP_x          5x

                      Blank Result                     10          10
                      CRQL                            5           5
                      Sample Results                  120          60
                      Final Sample Result              120          60

                   For both the "10x".and "5x" rules, sample results exceeded the
                   adjusted blank result  of  100 (or 10 x  10)  and 50 (or 5  x 10),
                   respectively.

-------
                                                        Procedure No.:  M2-VOA
                                                                    Revision: 1
                                                               Date: 06/30/1995
                                                                   Page 25 of 34
3.7 System Monitoring Compounds (Surrogate Spikes)

Laboratory performance on individual samples is established by means of spiking activities.
All samples  are spiked with system monitoring compounds (formerly referred to as
surrogates) prior to  sample purging,  the evaluation of the results  of these system
monitoring compounds is not necessarily straightforward.  The sample itself may produce
effects due to such factors as interferences and high conc^entratiohs of analytes.  Since the
effects of the sample matrix are frequently outside the control Sfpe^wl^bry and  may
present relatively unique problems, the evaluation and review of ^pPra^T'on specific
sample results is frequently subjective and demands analytical experience'and  professional
judgment. Accordingly, this section consists primarily  of guideUnes," in some cases with
several optional approaches suggested.                        •-*—>—^s.,.,

3.7.7  Acceptance Criteria

1.  Three system monitoring compounds (1,2^cMoroethane^4,'^rp^flubroyenieene, and
    toluene-d8) are added  to all samples and  blanks to measure  their recovery in
    environmental samples and blank matrices.            :    ;^1^X*-   '

2.  Recoveries for system monitoring compounds in volatile samples and blanks must be
    within the limits specified in Appendix A and the SOW.

3.7.2  Review Items

Form n VOA and chromatograms.

3.7.3  Evaluation Procedures

1.   Check chromatograms to verify the recoveries on the System Monitoring Compound
    Recovery Form—Form n VOA.

2.   The following should be determined from the System Monitoring Compound Recovery
    form(s):

    a.        If any system monitoring compound(s) in the volatile  fraction  is out of
             specification,  there should  be  a reanalysis to  confirm  that  the non-
             compliance  is due  to  sample  matrix  effects  rather than   laboratory
             deficiencies.                                       '

    NOTE:   When  there  are unacceptable system monitoring compound  recoveries
             followed by successful analyses, the laboratories are required to report only
             the successful run.

-------
                                                          Procedure No.:  M2-VOA
                                                                      Revision:  1
                                                                 Date: 06/30/1995
                                                                     Page 26 of 34
    b.       The laboratory failed to perform acceptably if system monitoring compounds
             are outside criteria with no evidence of re-analysis. Medium soils must first
             be re-extracted prior to, re-analysis when this occurs.

    c.       Verify that no  blanks  have system  monitoring compounds  outside  the
3.  Any time there'are^two or mpre analyses for a particular sample, the reviewer must
    determine which a^ the best data to report Considerations should include but are not
    limited to: .  ?-xฃij&;&&^-  '-'           :.   '    •.  .'•   .'.•'  Y'•*•.'•

    a.       System monitoring compound recovery (marginal versus gross deviation).

    b.       Technical/holding times.   ;_                 ...   ;
   -  •            •'•• .•='.'•&>-•.••'*•;.:*"'"'•••'•<•'' . '•   '     '  .    •        .-   •
                 -• ' -fiS-^vJ-•.:;.-• -.        '         .       .  .  • .'
    c.       Comparison of the values of the target compounds reported in each sample
                 • *	i  -:;.\,.-^i , '.-ป.!;v-_' ซv.'ป' ...   •        *^      . *       • *    - .  .           * •
    d.       Other QC information, such as performance of internal standards.
                    ."."'." -v-  ' *
3.7.4 Action

Data are qualified based on system monitoring compounds results if the recovery of any
volatile system monitoring compound is  out  of specification.   For system  monitoring
compound recoveries out of specification, the following approaches are suggested based on
a review of all data from the package, especially considering the apparent complexity of the
sample matrix.  (Also, see Table M2-VQA-6, below).

1.  If a system monitoring compound in the volatile sample has a recovery greater than the
    upper acceptance limit:                .

    a.       Detected volatile target compounds are qualified "J".

    b.       Results for non-detected volatile target compounds should be qualified "UJ"

2.  If a system monitoring compound in the volatile sample has a recovery greater than or
    equal to 10% but less than the lower acceptance limit:

    a.       Detected volatile target compounds are qualified "J".

    b.   ,    For non-detected volatile target compounds, the sample quantitation limit is
             qualified as approximated, "UJ".

-------
                                                         Procedure No.:  M2-VOA
                                                                      Revision:  1
                                                                Date:  06/30/1995
                                                                    Page 27 of 34
3.  If a system monitoring compound in a volatile sample shows less than 10% recovery:

    a.       Detected volatile compounds are qualified "L".       ._••••-'
4.
5.
6.
                                                                        4
-------
                                                        Procedure No.: M2-VOA
                                                                     Revision:  1
                                                                Date:  06/30/1995
                                                                   Page 28 of 34
9.  When both the initial analysis and the reanalysis have system monitoring compound
    recoveries outside of criteria, the data summary form should normally contain the
    highest  concentration obtained for each, compound  detected, provided that system
    monitoring compound recoveries in the analysis being reported do not suggest a high
    bias.  However, if a demonstrated laboratory contaminant is detected in one analysis
    but not in the other,  the negative result, may be more  appropriate to report
        •  ,-.  •;.-'.i.-r'->j.^-'--::'-ซ.v-.-.-   '•    '        '  .  •.••:••'•"•/>••&-.'-.S-";   -
    When the reanalysis of a sample is within the system monitoring compound recovery
    criteria, the laboratory is required to provide only data for the acceptable analysis. If
    both sets of data are provided, and if a compound was detected in the initial analysis
    but not in the reanalysis, then the positive result should be reported (provided the
    compound is not a demonstrated laboratory contaminant).  The reported result should
    be flagged as estimated 0, due to possible sample inhomogeneity.     .

3.8  Matrix Spike/Matrix Spike Duplicate

Data for matrix spike/matrix spike duplicates (MS/MSD) are generated to determine long-
term precision and accuracy of the analytical method oh various  matrices  and to
demonstrate  acceptable compound recovery  by  the  laboratory at  the time of sample
analysis.  These data alone cannot l>e used to evaluate the  precision  and accuracy of
individual samples. However, when exercising  professional judgment, this data should be
used in conjunction with  other available QC information.       .         - ^

3.8.1  Acceptance" Criteria

1.  Matrix spike (MS)  and  matrix  spike duplicate (MSD) samples are analyzed  at  a
    frequency of one MS and MSD per 20 samples of similar  matrix.

2.  Spike recoveries should be within the advisory limits provided on Form m VOA-1 and
    VOA-2.                              .

3.  Relative percent difference (RPD) between MS and MSD recoveries must be within
    the advisory limits provided on Form HI VOA-1 and VOA-2.

3.8.2  Review Items

Form III VOA-I and VOA-2, chromatograms.

3.8.3  Evaluation Procedures

1.  Verify that MS and  MSD samples were analyzed at  the required frequency  and that
    results are provided for each sample matrix.

-------
                                                        Procedure No.: M2-VOA
                                                                     Revision:  1
                                                               Date:  06/30/1995
                                                                   Page 29 of 34
 2.  Inspect results for the MS/MSD Recovery on Form m VOA-1 and VOA-2 and verify
    that the results for recovery and RPD are within the advisory limits.
3.  Check that the matrix spike recoveries and RPDs were calculated correctly.

4.  Compare %RSD results of non-spiked compounds between the original result, MS, and
    MSD.     .

3.8.4 Action

1.  No action is taken on MS/MSD data alone.  However, using informed professional
    judgment, the  data reviewer may use the MS-and MSD results in conjunction with
    other QC criteria to determine the need for some qualification of the data.

2.  The data reviewer should first try  to determine to what extent the results of the
    MS/MSD affect the associated data. This determination should be made with regard to
    the MS/MSD sample itself as well as specific analytes for all samples .associated with
    the MS/MSD.

3.  In those instances where it can be determined that the results of the MS/MSD affect
    only the sample spiked, then qualification should be limited to this- sample alone.
    However,  it may be determined through the MS/MSD results that  a laboratory is
    having a systematic problem in the analysis of one or more analytes, which affects all
    associated  samples.

4.  The reviewer must use professional judgment to determine the need for qualification of
    positive results of non-spiked compounds.

5.  When non-spiked compounds are present in either the MS or MSD  results, a table in
    the data review narrative is constructed showing original (unspiked)  sample results for
    non-spiked compounds, non-spiked compounds present in the MS and MSD  and the
    calculated  %RSD.

    NOTE:  If a field blank was used for the MS/MSD,  a statement to that effect  must be
            included on the ORDAS and noted for the TPO.

3.9  Internal Standards

Internal Standards (IS) performance criteria ensures  that  GC/MS sensitivity  and response
are stable during each analysis.

-------
                                                        Procedure No.:  M2-VOA
                                                                     Revision:  1
                                                               Date: 06/30/1995
                                                                   Page 30 of 34


3.9.1  Acceptance Criteria

1.  Internal standard area counts must not vary by more than a factor of two (-50% to
    +100%) from the associated calibration standard.

2.  The retention time of the internal standard must not vary more than ฑ30 seconds from
    the retention time of the associated calibration standard.
                             ;'•'•''     i

3.9.2  Data Requirements and Data Retrieval

Form YE VOA and chromatograms.

3.9.3  Evaluation Procedures

1.  Check the internal standard retention times and areas reported on the Internal Standard
    Area Summary (Form vm VOA).

2.  Verify that all retention times and IS areas are within criteria.

3.  If there are two analyses for a particular fraction, the reviewer must determine which
    are the best data to report.  Considerations should include:

    a.        Magnitude and direction of the IS area shift.

    b.        Magnitude and direction of the IS retention time shift.

    c.        Technical holding times.

    d.        Comparison of the values of the target compounds reported in each fraction.

    c.        Other QC.

3.9.4  Action

1.  If an IS  area count for a sample or blank is outside -50% or +100% of the area for
    associated standard, then:

    a.        Positive results for compounds quantitated using that IS should be qualified
             with'T'.                              .

    b.        Non-detected compounds quantitated using an, IS area count greater than .+
             100 % should be  qualified ' 'UJ''.

-------
                                                         Procedure No.:  M2-VQA
                                                                      Revision:  1
                                                                 Date:  06/30/1995
                                                                     Page 31 of 34-
    c.       Non-detected compounds quantitated using an IS area count less than 50%
             are reported as the associated sample quantitation limit and qualified with
             "UJ".                     .                ;

    d.       If extremely  low area counts are reported, or if performance exhibits a
             major abrupt drop-off then a severe loss of sensitivity is indicated.  Non-
             detected target compounds should then be qualified as unusable, "R".

2.  If an IS retention time varies by more than 30 seconds:

    The chromatographic profile for that sample must be examined to determine  if any
    false positives or negatives exist.  For shifts of a large magnitude, the reviewer may
    consider partial or total rejection of the datajor that sample fraction; Positive results
    should not need to be qualified as "R", if the mass spectral criteria are met.

3.  If the internal standards performance criteria are grossly exceeded, then this should be
    noted for TPO action in the  ORDAS.  Potential affects on the data resulting from
    unacceptable internal  standard  performance  should be noted  in  the  data  review
    narrative.

3.10 Reported  CRQLs

The objective is to ensure that  the Contract Required  Quantitation Limits (CRQLs) are
accurate.

3.10.1   Acceptance Criteria

1.  The adjustment of the CRQLs must be calculated according to  the correct equation to
    account for dilution and moisture content as appropriate.

3.10.2   Review Items

Form I VOA and chromatograms.

3.10.3   Evaluation Procedures

1.  Verify that  the CRQLs have been adjusted to reflect all  sample dilutions and  dry
    weight factors  that are not accounted for by the method.

3.10.4   Action

1,  If any discrepancies are found, the laboratory may be contacted by the  designated
    representative to obtain additional information that could resolve any differences.  If a

-------
                                                        Procedure No;: M2-VOA
                                                                     Revision:  1
                                                               Date:  06/30/1995
                                                                   Page 32 of 34
    discrepancy  remains  unresolved, the  reviewer must use professional judgment to
    decide which value is the best value.  Under these circumstances, the reviewer may
    determine qualification of data is warranted.  A description of the reasons for data
    qualification and the qualification that is applied to the data should be documented in
    the data review narrative and in the document support.            ,
          '  . • '•   .•_.•;•'..'•    '  •           •,-•";.-• •;•>;.^vjfe: •;••''• '•
2.  The reviewer must assure that any .results in error by more ;timn ^1,0 percent are
    identified and corrected on the sample data summary.  If laborBo^ resuljmission is
    not performed, the reviewer should document his/her changes  to;the data in the
    narrative and support; documentation. Calculation errors should also be noted on the
    ORDAS.

3.  Numerous or significant failures to properly^eyaluate and adju^MIJRQLs should be
    noted for TPO action  on the ORDAS.

3.11  Tentatively Identified Compounds

Chrpmatographic peaks in volatile fraction  analyses that are not target analytes,  system"
monitoring compounds or  internal standards are potential tentatively/identified compounds
(TICs).

3.11.1  Acceptance Criteria

For each sample, the laboratory must conduct a mass spectral search of the NIST library
and report the possible identity for the  10  largest volatile fraction peaks which are not
system monitoring compounds, internal standards, or target compounds, but which have an
area or height greater than 10 percent of the area or height of the nearest internal standard.
TIC results are reported  for  each sample on  the Organic Analyses Data Sheet (Form I
VOA-TIC).

    NOTE:   Since the SOW revision of October 1986, the CLP does not allow the
             laboratory to report as tentatively identified compounds any target compound
             which is properly reported in another fraction.  For example, late  eluting
             volatile target compounds should not be reported as semivolatile TICs.

3.11.2  Review Items

Form  I VOA-TIC chromatograms.

3.11.3  Evaluation Procedures

1.  Blank chromatograms should be examined to verify that TIC peaks present in samples
    are  not found in blanks.   When a low-level non-target compound that is a common

-------
                                                          Procedure No.:  M2-VOA
                                                                       Revision:  1
                                                                  Date:  06/30/1995
                                                                      Page 33 of 34
    artifact or laboratory contaminant is detected in a sample, a thorough check of blank
    chromatograms may require looking for peaks which are less than  10 percent of the
    internal standard height, but present in the blank chromatogram at  a similar relative
    retention time.

2.  The reviewer should be aware of common laboratory artifacts/contaminants and their
    sources  (e.g.,  aldol  -condensation products, solvent  preservatives,  and  reagent
    contaminants).  These may be present in blanks and not reported 'as sample TICs.
    Examples:
             Common laboratory contaminants:   COj (m/z 44), siloxanes (m/z 73),
             diethyl ether, hexane, certain freonji (1, l,2-trichloro-l,2,2-trMuoroethaneor
             fluorotrichloromethane), and phthalates at levels less than 100 ug/L or 4000
                        '                 "''
    b.       Solvent preservatives such as  cyclohexene which is a methylene chloride
             preservative.  Related by-products include cyclohexanone, cyclohexenone,
             cyclohexanol, cyclohexenol, chlorocyclohexene, and chlorocyclohexanol.
                        . • •   \~
    c.       Aldol  condensation  reaction products of  acetone include:   4-hydroxy-4-
             methyl-2-pentanone,   4-methyl-2-penten-2-one,  and  5,5rdimethyl-2(5H)-
             furanone.                                  •           ""-

3.  Occasionally, a target compound may be identified in the proper analytical fraction by
    non-target library search procedures, even though it was not found on the quantitation
    list.  If the total area quantitation method was used,  the reviewer should request that
    the laboratory recalculate the result using the proper quantitation ion.  In addition, the
    reviewer should evaluate  other sample  chromatograms and check library  reference
    retention times on quantitation lists to determine whether the false negative result is an
    isolated occurrence or whether additional data may be affected.

4.  Target  compounds could be  identified in more than one  fraction.   Verify  that
    quantitation is made from the proper fraction.

5.  TIC concentration should be estimated assuming a RRF of 1.0.

3.11.4 Action

1.  All TIC results should be qualified "J", estimated  concentration,  on the laboratory
    Form I-TICs. '

2.  General actions related to the review of TIC results are as follows:

-------
                                                         Procedure No.:  M2-VOA
                                                                      Revision:  1
                                                                 Date:  06/30/1995
                                                                    Page 34 of 34
    If all contractually  required peaks were  not  library  searched and quantitated,  the
    designated representative could request these data from the laboratory.

3.  Blank Results

    Form I-T1C which contains sample results that are questioned by laboratory results,
    should be flagged "B" and a line drawn through these data for; emphasis (initialed and
    dated), on the Form MIC that is included in the validation .iqpi^)ง^.;••:',

    To be considered questionable, a sample TIC concentration must bb~ within 10 times
    the concentration of one of the blank results.  If different volumes/weights are used,
    the total amount of compound in the  extract  must be compared for sample versus
    blank.    For  VGA data,  an  instrument  level comparison! is ฃ -used  unless  the
    contamination    is.  proven  to   originate  during.  sample ."storage   (before
    preparation/analysis).  In general, blanks analyzed within the same case, by the same
    lab, may be cross-applied to either soil or water samples extracted or analyzed on
    other days.                    '...•'•               .'•.''•  '
                                                              ' • • j.     '. • •*
    To question a sample result, only presumptive evidence for the presence of the
    compound in the blank is necessary.  The presence  of  the TIC  in the blank is
    suggested in any of the following situations:                       ,

    All blank results must be attached in the  support documentation  section of the data
    review.

4.  When a compound is not found in any blanks, but is a suspected artifact of common
    laboratory contaminant,  the  result may be qualified as unusable, "R", and a  line
    drawn through the result (initialed and dated) on a copy of the Form I-TIC that is
    included in the validation report.                                        .

5.  Physical constants, such as boiling point, may  be factored into professional judgment
    of TIC results.

6.  Failure to properly evaluate and report TICs should be noted  for TPO action on the
    ORDAS form.

-------
                          Appendix B
        Validation of Semivolatile X)rganic Analyte Data

                 • Subappendix B-l covers Level Ml

                 • Subappendix A-2 covers Level M2
WDCR565/03i52

-------
                 Appendix B

Validation of SemivolatUe t>rganic Analyte Data
'"••'•-  "	mj^&c&^:-<:-'V.-.--"•••••• -:-;:..?.*v-: •" • •• • •
        j-=t^^^?^-C'--:-- •-'••; ••  ' •' - v:.V- -
        • Subappendix B-l covers Level Ml


        * Subappendix A-2 covers Level M2

-------
                                                      Procedure No.:  Ml-SVOA
                                                                    Revision:  1
                                                               Date:  06/30/1995
                                                                   Page 1 of 19
                                 Appendix B-l
          Validation of Semivolatile Organic Analyte Data
                            Manual Level Ml
                       1; Purpose and Applicability
                        if ••.-••      •     •      -   -             •     .
This procedure provides ^t^-byrstep instructions to manually validate  the semivolatile
organic analyte (SVOAJ^iSati^usiiSg'ihe manual innovative data validation approach at Level
 m'   •    '  •  '"   •: "\    ' J ''                                          '
    '     ''    •
         .    .                                                  ....
Hie procedure is appli^lejipithersyOA datax>btained using the Contract Laboratory
Program Statement of^ori^CILP^SOW).  Harcfcopy data conforming to the CLP SOW
specifications :anฃessieirafi^

Data validated .(iusmg;^^l||re^edpte are considered  usable for the  following types of
purposes;  however, ^tn^|;:^^;must decide  on a case-by-case basis  whether the
procedure is suitable for their intended data uses. The suggested data uses are:

      •     Oversight of activities led by other parties

      •     Comparison to action levels

      •,    Initial site investigation

      •     Contamination sources


                  2. Quality Control Measures Checked

Table Ml-SVOA-1 highlights the quality control (QC) indicators evaluated under this data
validation procedure.


                                3.  Procedure

The following subsections  describe for  each of the QC indicators the acceptance  criteria,
location and retrieval of QC data, evaluation of the  QC data, actions taken in the event the
QC acceptance  criteria are exceeded, and documentation of the QC  violations  in  a
standardized report form.
                                                        U.S. EPA Headquarters Library
                                                              Mail code 3201
                                                        1200 Pennsylvania Avenue NW
                                                          Washington  DC 20460

-------
                                                    Procedure No.:  Ml-SVOA
                                                                 Revision:  1
                                                            Date: 06/30/1995
                                                                 Page 2 of 19
                             Table Ml-SVOA-1
                     QC CHECKLIST FOR LEVEL Ml
                   CLP RAS SEMIVOLATILE ORGANICS
             QC MEASURES
. Action Level. Notification
Instnunent Tune
                                                      Manual
Initial Calibration (RRF)
Initial Calibration (%RSD)

 Continuing Calibration (RRF)
 Continuing-Calibration (%D)
Laboratory Blank
Surrogate Recovery
Dilution Factor
Moisture Content
Mass Spectra
Chromatograms
Raw Data

-------
                                                        Procedure No.: Ml-SVOA
                                                                      Revision:  1
                                                                Date:  06/30/1995
                                                                     Page 3 of 19
Reporting requirements for Level Ml are:

       •     Hand annotate the Form I's, including
                    Data validation qualifiers
             -      Sample identification number
                    Sampling location

       •     Provide a narrative that includes
                    A statement that defines the level of the data review, i.e., Ml
                    Major problems associated with analysis
                      -            •              **-                 .        -
       •     Include the following attachments
             -      list of data validation qualifiers
                    Support documentation including forms that support assigning data
                    qualifiers     -   "
                    Chain  of custody form

3.1  Action Level Notification

The purpose behind action level notification is to make the EPA Remedial Project Officer
(RPM) or the Site Project Officer (SPO) aware of the potential human health risk at the
site.  Li accordance with the Region m Hazardous waste division policy, the EPA RPM or
SPO must be promptly notified of any contaminant exceeding the established action level or
the 10-day health advisory limit.  The data for contaminants exceeding the action levels
must be validated as a top priority and reported to the RPM or SPO.  Validation of the rest
of the data may then be completed normally.

3.1.1  Acceptance Criteria

EPA's Office of Solid Waste and Emergency Response has established  10-day advisory
limits or the action levels for several organic compounds and elements of special health risk
concerns based on the Safe Drinking Water Act. The semivolatile organic compounds and
their 10-day  health advisory limits apply only to aqueous samples and are listed in Table
Ml-SVOA-2. The criteria for action level notification are as follows:

       • '     The contaminant concentration must be equal to or above the established 10-
             day health advisory limits.

       •     Data for contaminants exceeding the action levels must be validated as a top
             priority.

-------
                                          Procedure No.:  Ml-SVOA
                                                       Revision:  1
                                                  Date: 06/30/1995
                                                       Page 4 of 19
The  following  EPA personnel must  be  notified  of the action  level
exceedances:

      EPARPMorSPO

      EPA Section Chiefs:
      -     Site Investigation (SI)

      -     Remedial

      —     Enforcement

      -     RCRA

      EPA Section lexicologists:

      —     Enforcement

      —     Superfund

             RCRA

The remaining data validation should be completed per normal procedures.

Any  special  instructions  from the Hazardous Waste Division should  be
followed.

Records should be kept of the data review, action level notification and any
follow up instructions and actions.
Table Ml-SVOA-2
SEMIVOLATILE ORGANIC ANALYTES AND ACTION LEVELS
Compound
1 ,3-Dichlorobenzene
Pentachlorophenol
Action
Level*
8,930
300
Compound
1 ,4-Dichlorobenzene
—
Action
Level*
10,700
• . —
*A11 units are ug/1.

-------
                                                        Procedure No.r Ml-SVOA
                                                                      Revision:  1
                                                                Date:  06/30/1995
                                                                     PageS of 19
3.1.2  Review Items
All data required to perform Level Ml validation, as detailed in the following sections are
necessary for carrying out action  level  notification.  The location of the data and their
retrieval procedures are also discussed below.

3.1.3 Evaluation  Procedure

The evaluation process preceding action 'level notification  will primarily  consist  of
comparing the results on Form I's with the action levels presented in Table Ml-SVOA-2.
Following the identification of the contaminants exceeding the action levels, focused data
validation should1 be  performed1 using  the criteria, and procedures  described  in  the
appropriate se(jtionsi>elow; '  ff H           •*•."•--

3.L4 Action               -                   ;    •     "  •    .   •  "  ' •    •;

The action resulting from focused: data validation will be the  notification of action level
exceedance to the personnel identified above in Section 3.1.1.    '

3.1.5 Reporting    .

Copies of Form I's can be used to highlight the contaminants above the action levels. The
findings of the focused validation can be summarized in a memorandum, and the data
qualifiers resulting from focused validation may be written on the Form I's.  The marked
up  forms should  be clarified that they represent  validation  of only the contaminants
exceeding the action levels, and not all data.

3.2 Evaluation of Mass Spectra for the Detected Compounds

The primary  QC  indicator checked  in Level  Ml is the mass spectra  for  the detected
compounds.  This indicator pertains to evaluating a compound's presence by matching its
mass spectrum with a standard (known) mass spectrum for the compound. No assessment
is  made of the reported quantity  of the compound or any quantitative quality control
indicators that could lend some uncertainty to the reported value.

3.2.7  Acceptance Criteria

The acceptance criteria for  mass spectral  matching are given in the  CLP Functional
Guidelines and are as follows:

       •      All ions present in the standard (known)  spectrum  at a relative intensity
            ; greater than 10 percent must be present in the sample mass spectrum

-------
                                                        Procedure No.:  Ml-SVOA
                                                                     Revision:  1
                                                                Date:  06/30/1995
                                                                     Page 6 of 19
             The relative intensities of the qualifying ions (those above 10 percent relative
             intensity) must .be within +/-20 percent between the standard and  sample
             spectra, for example, an ion with a relative abundance of SO percent in the
             standard spectrum must be present between a relative abundance of 30 and
             70 percent in the sample spectrum)                ;

             Ions, present in the  sample above 10 percent relative abundance must  be
                  *               •  *            *            . .ป "• ?-.*^ '-t-'X'^'--,- ''.
             accounted for.     -  •          '       .;.-.-  •  . : ':'.:!:..-  '•• •• :
3.2.2  Review Items
                                     .
Form I's for each field sample, field blanks, and laboratory blanks mduded in a Sample
Delivery Group (SDG) are necessary to compiled list of the detected compounds.  Mass
spectra are then necessary, for each of the detected compounds in a sample.-The required
mass spectra include both the sample spectra as well as the standard spectra,  r
With respect to the blanks, it may be beneficial to evaluate the blanks before, the sample
mass spectra are evaluated,  If a compound is found to be a common contaminant, i.e., it *
is present in any one of the blanks and also in the sample, the concentration in the sample
should be  evaluated -with respect to the highest blank contamination using the 5 (or 10)
times  criteria before proceeding  with the elaborate mass spectral evaluations. .The effort
required for  the  sample mass  spectral  evaluation  may be  significantly-, reduced  by
performing the blank evaluation first.  The mass spectra for the blanks; however, should be
evaluated to ensure that the contaminants were identified properly.

All sample Form I's are generally located together  in front of the data package. Copies of
the Form I's can also be found in the .detailed sample data package organized by sample;
i.e., the Form I and its substantiating raw data for  a sample are placed together.  The raw
data include a quantitation report and mass spectra for each detected compounds in that
sample are.   A standard (known) mass spectrum,  and two  sample mass  spectra (one
unaltered and another background-subtracted) are generally provided.  The presentation of
the mass spectra differs for different instrument manufacturers, but information necessary
for evaluating mass spectra is always provided in  some form.  Hands-on experience with
different brands of  mass  spectrometers,  although  not  essential,  can  be helpful  in
interpreting the information.

Copies should  be made of the Form I's, preferably from the summary data package, if
included.  Otherwise, the forms can be pulled out from the raw sample data package. All
detects should then be highlighted with a marker or other convenient means.  The raw data
should then be tagged for every sample  for evaluating the  mass spectra  and the
chromatograms (reconstructed ion current profiles).

-------
                                                        Procedure No.:  Ml-SVOA
                                                                      Revision:  1
                                                                 Date:  06/30/1995
                                                                     Page 7 of 19
3.2.3  Evaluation of Mass Spectra

Compare the highlighted Form I's with the tagged raw sample data.  Verify that raw data
are provided for every sample, and a Form I exists for every sample represented by raw
data.  Verify on a sample-by-sample basis that mass spectra are provided for every detected
(highlighted) compound reported on the Form I's.  A cross verification  should then be
performed that all target compounds for which mass spectra are provided are also reported
as detected on Form I's. ^; •.?&&'••   '  " '   , :.-•.'•<  v''
Mass spectral comparison  routines based on  the  evaluation criteria presented earlier
(presence of major ions and their relative intensities) are built into the instrument software;
therefore, performed automatically? A listing of ions and their relative intensities is not
produced; rather a score based on the extent of^the fit of all criteria is produced. Hie
score is based on a basis of 100 or 1000 for a perfect match of the presence of major ions
and thek relative abundances^ f^e score is printed on the quantitation report as a "q"
value, a score upward of 60 percent of the maximum (upward of 60 or 600) is generally
considered acceptable for positive identification for a compound. <  .  :

Visual comparison of a mass spectrum involves looking for the base ion (the mass fragment
with the highest intensity),  the parent ion (mass fragment equal to the molecular weight of
the compound) and other characteristic ions representing removal of one or more functional
groups (such as -CH3, -CH2-, -Cl, -OH, or a combination thereof) from the-parent ion, the
base ion or other ions.  The removal of functional groups is often successive and produces
a fingerprint pattern for a type of compound; i.e., straight-chain hydrocarbons produce a
characteristic envelope of mass fragments separated by 14 mass units. Similarly, aromatic
compounds produce a peculiar fingerprint.  The base ion represents the most stable mass
fragment; therefore, it is always present. However, a parent ion often is unstable and may
not exist at all or may exist at a low abundance or relative intensity.

Compare the standard (known) mass spectrum  with  the . background-subtracted  sample
spectrum for the presence of the base, parent  and  other  characteristic ions.  Although
background-subtracted and  unaltered mass spectra are generally provided, the former type
of spectra are much cleaner looking due to the subtraction of column bleed or other broad-
based interferences;  therefore, better suited for comparison.  Generally,  the presence of
major ions and overall matching  of the fingerprint pattern between the  standard and the
sample. spectra can be considered satisfactory.  The comparison is rather subjectivcj and
requires a trained-eye to deduce the information.
                    •\
Interference still may be present in a background-subtracted mass spectrum due to coeluting
compounds (as  compared  to column bleed or  broad-based interferences).   Unless the
interfering compound is an  isomer or an analog  of the target compound in question, the
fingerprints produced by the target compound and the interfering compounds can be quite

-------
                                                        Procedure No.:  Ml-SVOA
                                                                      Revision: 1
                                                                 Date: 06/30/1995
                                                                      Page 8 of 19
different.  Inseparable isomers reflect a limitation of the gas chromatpgraphy without any
recourse.  Concentrations for such isomers may be reported either as all contribution from
one isomer or the value may be divided equally between the two isomers.  For example,
difficulties are often seen in separating benzo(b)fluoranthene and benzo(k)fluoranthene by
gas chromatography.. These compounds also produce similar mass spectra; therefore, it is
quite difficult to; tell -them apart   Homologous  compounds  generally have  different
retention times; therefore, may not be of much concern. Thus, coeluting .compounds do
not pose significant jirpblems .when evaluating the mass spectra.
                                '            '              '
Sddom does a CH^ laboratory incorrectly  interpret a mass spectrum, but there  is a
subjective element  to mass  spectral  interpretation.   While working within the CLP
guidelines, the laboratories ^do have their own reporting practices at the limit  of the
mstmment's senativity.^Most of the mass spectral identification problems occur near the
limit of detecticravwhere;the differences in the relative intensities are not easily discernible.
Special attenticm should be paid in evaluating mass spectra for compounds detected at low
•9ฐnCentrat^   -:'.•••    .     '     "''.'•'..-••'.    '•  •
Rearrangements and, other side-reactions often occur inside a mass spectrometer.  These'
phenomena produce rnass fragments that are  not easily accountable from the structure of
the parent compound; WhUe it is desirable that a data validator possess the knowledge to
interpret complex  mass spectra, . for the  routine CLP  analyses,  such  expertise is
unnecessary.

3.2.4 Action

In the event the  mass spectrum of a detected  compound does not at all resemble the
Standard spectrum or has extremely poor matching, the compound should be considered
undetected.  In this case the detected value should be changed to the CRDL for that sample
and  flagged as undetected,  "U."   (NOTE:   This  action is  in contrast to the EPA
Functional Guidelines which recommend rejection of data.  It is felt  that an outright
rejection of data is not justified.  If the mass spectrum produced at a  target  compound's
retention time does not match the standard spectrum, a conclusion can be drawn that the
target compound is not present and the mass  spectrum may be due to something else. In
such a case the target compound should be considered as undetected and data should not be
rejected.  Make sure that the target compound in question does produce a good mass
spectrum by inspecting the standard mass  spectrum from the calibration and the fit score on
the quantitation report for  the daily or continuing  calibration.   Data for undetected
compounds are usable for many purposes such as risk assessment; therefore, rejection of
data is not beneficial to a project.  The conflict of improper characterization should be
brought up and rectified with the laboratory.)

If there is some evidence of the compound's presence (as determined by visual matching of
the base ion, parent ion and fingerprint  pattern  despite a poor  fit score), the compound

-------
                                                        Procedure No.:  Ml-SVOA
                                                                     Revision:  1
                                                                Date:  06/30/1995
                                                                     Page 9 of 19
should be considered tentatively identified.  The reported value should not be altered in this
case; however, a data qualifier code, "N", should be appended to the data point to denote
tentative identification.                     •

In both of the above cases, evaluation of additional information such as the retention time
and frequency of detection in other field samples is warranted. Previous site history is also
an important evaluation; .however," information essential to perform such an evaluation (in
contrast to  'validation') may not be easily available to a data validator.  Nonetheless, it is
incumbent  uppn a data.validator-to make appropriate recommendations to 4he project
manager or the; remedial^/officer.^

3.2,5  R&orfag  i^ฑ^~:f:   •/;'••'    •   ;  '".••;.'.  ';J-'••:;:~;.?'  ;"  ;'   ,

The most convenient fonn of reporting the mass spectral evaluation is through a brief
memorandum to appropriate authorities.. The highlighted Forms Is and the mass spectra in
question should be appOTd6d tb/me memorandum in support of tiie  conclusions.  Also fill
out and attach the Mass Spectral Evaluation Form,  as depicted in Table Ml-SVOA-3, to
the report as a record of v what was done.  Record the date(s) of analyses.  Field sample
numbers should be transcdbed from the chain-of-custody in the Sample Identifier column.
Notations-may be placed under the "MS" column  for each sample to describe the mass
spectral evaluation.  The following notations are suggested:

       •     X—Acceptable mass spectrum.

       •     N—Tentative identification.  Partial proof of a compound's presence, but all
             identification criteria not met.

       •     U—No mass spectral match. Using professional judgment, the compound is
             considered undetected.

       •     R^-No mass spectral match. Using professional judgment, the compound
             data are rejected from further use..

3.3  Evaluation of Chromatograms

The purpose behind .evaluating the chromatograms is to get an idea regarding potential false
negatives, and gross analytical errors.  Checking for positive data as described under mass
spectral evaluation does not offer any insight into data that are not reported (i.e., reported
as nondetects).  Laboratory error or gross interference from other compounds could be the
reasons for erroneous reporting.  The gas. chromatograms or the reconstructed ion current
(RIC) profiles are the primary too! used for the false negative evaluation under Level Ml.

-------
          Case No.:.
          SDG No.:.
Data Reviewer:	
         Site:	
                                     Procedure No.:  Ml-SVOA
                                                   Revision:  1
                                              Date: 06/30/1995
        Table Ml-SVOA-3,
SVGA Mass Spectral Eraloation
                             Summar
Dates Analyzed:
lot. ID:
Sample Identifier
1. " •
2. ' : •
3.
4. ' •'."'. ': • '
s.-1 •-•' '' ' ' ' ''•• '• '.'•'
•6. '•' '' "-.'.' '• •'• • -•"••
1. '•'" . '-' -:' ' '";' ••' ' "
'•ซ.'• ' ' • "-•
9. ' '•' S • ••:••'••'
10. ;
SPCC <*) CCC(*|
"
Addk(A)
Compound: • ' Bam/Neutral (BN)
Phenol • • • - f*VA'
bUO-CUnoedivnedier VfiN)
2-CMoraohenol ' fAI
1.3-Diehlorooenzeoe ' fBNl
1. 4-Diehlorobenzene fMBN)
1.2-Diehlurabeuuiie fBN
2-MetfivlDhenol . fAI
2 r-oxvfaiifl-CMonปRmama fBJfl
4-Metbvbhenol fAI
N Nhroio-di-o-DroovUnune fBNVft
HexacbJoroethane • fBNl
Nitrobenzene fBW
Isrmhorone fBNl
2-NitroDhenol fAV*1
2.4-DimethvIohenol CA1
bis(2-Chloroethoxv>methane (BN)
2.4-DichloroohenoI fAI
1.2.4-Trichlorobenzene (BN)
Naohthalene CBN}
4-OhloroaniIine fBN)
Hexachlorobutadiene fBN)f*)
4-Chloro-3-methvlDhenol (M(*)

Hexachlorocvclooentadiene CBNXtf)
2 4 6-Trichloroohenol f AV*
2 4 5-TrichloroohenoI f A)
2-Chloronaohthalene fBN*
2-Nitroaniline (BNV
Dimethvlohthalate fBN)
Acenaohthvlene fBhO
2.6-Dinitrotoluene fBN)
3-Nitroaniline fBN)
' Acenaohthene f*1fBN)
2 4-DinitroohenoI (M(fT>
4-Nilrophenol (AW/SO
Commenta on Chromatognmi
1.
2. ' •
3. • •' ' • : •
4. .....' •• - •
5.' .'-; •• . •' ' . •-••'•"' '• '.-:'.: . -, . • ' ' . .
*v •••• • • : ; ••'••' '.".' . ••" ' •• • •'•'• •. • ••-.. ••_
7. ' V- •••• . "• . ••• • •. '• ••••. .
8. '"'• ' ; •• "• • • ' .'• ' •"••' .. -:'::. ' . >
9. •'••'' .. ' ' • • •'• '•'•': ' •
10.. . -. • :••'•••' .'.'•••:'•:"..•' .''• ' .
.'.. .'• -..•'.-•..•,•-'-. , - .
• ••ir-.'.
MS



































RT



































- ' 2 ••:'••
MS



































RT



































.',: 3 " '
MS

".
••













^


















RT





•





























•• •' ^ 'Samnte Identifier "••••,•..' • •• ••"•':'. .^: '•• '; ".'•":"
--. -4 •••
MS


•


i





























RT
. =


































;.;• V .
'MS





















ฐ




•- . '








RT



































6
MS



































RT



































":**• •
MS



































RT








• .-


























•' '•%:'••••
MS
t '


/



--,
-o


























RT







.



























'•' '• 9 ' '
MS



































RT




















I














id ;
MS



































RT

-. ..

- .
















ซ'















-------
Procedure No.:  Ml-SVOA
            Revision: 1
        Date:  06/30/1995
srcc an ccc(*
Acidic (A)
Compound: Base/Neutral (BN


DiethvbihthiUtfi fBNl
' 4-ChkHnnlienvl-ohenvIedier (tttf\
FhlOfNID • " (tttt
4-Ntonanttliwt • • • (BN1
. 4.6-D5nhro-2-mdhvtehenol (A1
N-NKrotod'roheflvlmnine V fBNSf*)
4-BrornoohenvJ-ohenvleflier ^(Bffi

RefltocUoroohenol . f AV*I


• CWnzale "'"'•'• '"• nhdปlate ' fBNI
Chraene fBNI
DMKOctVปnปi*ii1irte /BfnC*1
• PcinofbVfliioBMithenfl ' (Eti
Bcmofldfluorafitfaen& •• fBN^
BefnofB^DVKfw ffihnr''
. IndenoH .2.3-crfhjvrene fBNI
Dibcnz^A fi^anthracene ^BN
Benzofe.h.rh)ervlene fBN)
Sample Identifier
1
MS







.. .' .:
iS"?t****

'•• -: ':-•-



.'•^-•;7





'* -X;








RT








<&Ki:.
.-'if-.
.(•,.•;•

. .ซ.:•
• ''•:'•
' ?~'.:~t '

~ -












1
MS






'
V;-
,~ •.'"*•-'*'
.-.'T-..'
..•••:.',:•. •

..-. .

*.- -
• -, 1-













RT








..-,•--*•
:' •:•- "




• .' y
.-.



'• .-•

N'-







3
"MS








' -•










• '









RT





























4
MS



•

























RT


" .











""















MS













•
IT "














5
RT





























6
MS





























RT






























MS









.•• •



















1
RT




• ' -,










' •














MS
























~~-~
--.



8
RT





























9
MS





•























RT





























10
MS


















ป •
*









RT
















-r









;



-------
                                                        Procedure No.: Ml-SVOA
                                                                    .  Revision:  1
                                                                Date:  06/30/1995
                                                                    Page 12 of 19
3.3.1  Acceptance Criteria

There are no EPA-established criteria for evaluating chromatograms for false negatives.
The criteria used for evaluating chromatograms are based on good laboratory and scientific
practices, and these are not hard and fast requirements. The suggested evaluation criteria
are:     ' •'  '. ",... .  ,...:-. 'i..,    .  ' .  • '    -, .      • .ซ.• .-. •. .'ฃ•'•'• .'•••'
             There should not be any significant peaks in me (Aroin^grams th^ are not
             accounted for as TCLs or  TICs.   Significant peaks are. those with a
             minimum peak height of 10 percent of the (dosest internal standard.

             The  chromatograms  should ' ideally  have.. base-line resolution  between
             adjacent peaks.  Also, mere should nobbe broad (unresolved) envelops in the
             chromatograms.                     '  -:-    •-:'/r    ""' ......
              '•-   ; - , :-.  '.:;,-  - . •    .  • '.-' :."-.-•' '.••:•->.<•-; K!
             There should not be abrupt shifts in the baseline.

             There should not be peak tailing or sharp rise in the peak fronts.
3.3.2  Review Items

Chromatograms, or the RIG profiles for each sample are necessary for the evaluation.  The
RIC profiles can be found in the front of the raw data package for  each sample.   The
quantitation report for each sample is also necessary to retrieve either the retention times or
information regarding scan numbers to compare with the peaks on the chromatograms.

3.3.3  Evaluation of Chromatograms

Visually inspect the chromatograms for all peaks  that appear to be  at least  10 percent
height of the nearest internal standard. Compare that the scan number or the retention time
that appears on the x-axis of the RIC profile with that listed on the quantitation report for
the TCLs or the library search record for the TICs.  Make sure that all significant peaks
are accounted for.

Also observe  the RIC  profile for peak resolution between adjacent peaks.   Poor peak-to-
peak resolution is indicative of degrading performance of the gas chromatographic column.
The values obtained from a  degrading system are prone to be inaccurate.   Obviously
certain isomeric or homologous compounds are difficult to separate.  But generally, there
should be at least 90 percent valley between the neighboring peaks.

-------
                                                         Procedure No.: Ml-SVOA
                                                                        Revision:  1
                                                                  Date:  06/30/1995
                                                                      Page 13 of 19
Inspect the RIC profile for broad, unresolved envelops.  These are generally indicative of
outside interference from a series  of homologous compounds  such as  straight-chain
hydrocarbons.  Especially, assess the interference with the internal standards and surrogates
under the envelop using the expected area counts.  The standard values that are far from
the expected values may be indicative of potential problems with the TCL detection or
quantification.  "
Inspect the RIC profile for abrupt shifts in the baseline. /Such shifts >are indicative of
problems with'instrument sensitivity or leakage in the system.  The area counts obtained
from shifted baseline are inaccurate, or even the detection of a TCL at low concentration
may be missed/     ;
     • V  .  :. . • .   ..  , '„ ' .,._.. -   •  • '-        '     '  '      -'             ,-
      •'•.•• .'• "."."•  A ..- "'"'••'>t  ' , '   -           -         '      . ' .    *   '-.•"..
Rapid peak rising or peak tailing indicate problems* with the gas chromatographic column,
such as depleted stationary phase on the column, decomposition of the stationary, phase or
creation of active sites. Again, a visual inspection of the RIC profile will yield information
on the shape ofthepeafc;^    .•'•<••'•    ••   •'•   '.     '   '/v-';.-^--^''/'--... "• '.:"•'    -.   '>•

3.3.4  Action  •        vv';'"-  .'. .  ..    .         :  .  --^ •:•':--•> •":; ;„•/,;.;'  •  -. *

Professional discretion must be. used when evaluating and  qualifying data based  on the
chromatographic evaluations.  An experienced chemist can generally infer the magnitude
and the frequency of the problem from the RIC profile.  If the probleitr appears to be
systematic, then it  should be brought  to  the  laboratories  attention  and resolved.
Intermittent problems may or may not require any action.  The following guidelines are
suggested when acting on RIC profile observations: -

       •      Any unaccounted TCL  peak with area equivalent to or  greater than the
              lowest reportable limit for the sample must be brought  to the laboratory's
              attention and resolved. Any unaccounted non-TCL (i.e., TIC) peak with an
              area equal to or greater than 10'percent area of the nearest internal standard
              must also be resolved with the laboratory.  TICs with less than 10 percent
              area are  not required to be reported according to the  CLP-RAS protocols.
              In the event,  the discrepancy cannot be resolved with the laboratory, the
              problems should be documented and brought to the attention of the CLP-
              TPO, the RPM and the SM.   The data for unreported TCL or TICs may be
              considered  unusable until the problems are resolved.

       •      If a peak resolution problem  is evident for  the samples, and appears to be
              systematic  (i.e.,  present in all calibration samples, QC  samples, and field
              samples and increasing as the run progresses,  additional QC measures such
              as the continuing calibration percent difference (%D),  and internal  and
              surrogate standard recoveries  in the vicinity of the affected peaks should be

-------
                                                        Procedure No.: Ml-SVOA
                                                                      Revision:  1
                                                                Date:  06/30/1995
                                                                    Page 14 of 19
             evaluated to determine if the peak resolution problem could affect detection
             or quantification.  If determined so, the positive data may be qualified as
             estimated, "J." Negative data may also be qualified as estimated, "UJ" if
             the ability to detect at low concentrations is also deemed to be jeopardized
             by poor resolution of adjacent peaks.    -
             Broad   envelops   of  homologous  .compounds  could:  interfere  with
             quantification ,or;even detection.  If the mteaference is evident from the
             recoveries of the internal and surrogate standards in the .vicinity of the
             envelop, associated.compounds may  also be  interfered,with.   Using
             professional discretion, the positive arid negative data may ^considered as
             estimated, 'T'^and^W', respectively.  If the project objectives cannot be
             met with the estimated  data, alternative sample preparation \and cleanup
             procedures-may: need to be developed and specified.  ';. v
       •      Discrete;  shifts'in'the:: baseline in the rmddle of a nm are m^      of
             intermittent problems. If the shift is due to leakage or change in the system
             pressure, me positive as well as negative data' may .be; considered estimated
             ("J"  and "UJ,  respectively).  The problem could be also due to some
             fluctuation in the instrument electronics which may lead to drastic changes in
             the sensitivity of the instrument to defect  the compounds.   As a note,
             professional judgment should be exercised in determining the severity of the
             problem. For example, the magnitude of a  drop in the baseline below the
             zero line may not be estimated and could be very significant. On the other
             hand,  a  drop that  yields a baseline  still  above zero  can be  put in a
             perspective with  the original baseline and a general appearance of the entire
             RIC profile.

       •      The problems with peak symmetry are indicative of system degradation, and
             should be brought to the attention of the  laboratory for a corrective action.
             Professional judgment should be used when and if qualifying any data due to
             unsymmetrical peaks.  First the problem should be defined in terms of
             persistence  throughout the chromatogram and also from sample to sample.
             Additionally,  the shapes and  area  counts for  the internal and  surrogate
             standards should be  evaluated  to see if the problem  could have affected
             compound  detection and/or quantification.    Data qualification  may  be
             uncalled for if the standard area counts are acceptable.

3.3.5  Reporting

The Mass Spectral Evaluation Form (Table Ml-SVOA-3)  may be used to note any brief
comments on the chromatographic evaluations. The comments may be noted against each
sample identifier.  For more descriptive comments, a separate sheet may be used.

-------
                                                        Procedure No.:  Ml-SVOA
                                                                     Revision:  1
                                                                Date: 06/30/1995
                                                                    Page 15 of 19
3.4 Evaluation of Retention Times

While not unequivocal in identifying a compound, the retention times are quite helpful in
confirming the presence of a detected compound.   Matching of mass spectrum and
retention time of sample with those of a standard yields higher credibility and confidence
level to the detection m the sample.  On the other hand, not matching the retention time
may or may not invalidate tiie: detection.  If a mass spectral match is made beyond any
doubt, poor matchirig of the retention time may not have adverse impact on the detection.
If mass spectral matching; is unacceptable or only partial, and the retention times do not
match, then a strong doubt can be cast on a compound's presence.

3.4.1  Acceptance Criteria               *-.,.-
              •  "  ! '••• >;'.;-: '••"•>::'•'ป:''  '' •  •.       "         ''.
The criteria for retention tim^               the EPA's functional guidelines as follows:

      The relative retention times (RRTs) must be within +/-0.06 RRT units of    ,  ;
      the appUcable internal standard RRT.                         -
               •    "   .'     '  '   -.           •               • •
3.4.2  Review Items

Quantitation  reports  for the sample and continuing  calibration are required  for the
evaluation.  These are located in the raw data for samples and standards.  A copy of the
continuing calibration quantitation reports may be  made or the reports pulled out from the
raw data to facilitate a comparison with the sample quantitation reports.

3.4.3  Evaluation of Retention Times

For the detected compounds, determine the relative retention time for  the compounds by
dividing their retention times with the retention time of their associated internal standard in
the samples as well as in the applicable continuing (or initial) calibrations.   The sample
RRTs must fall in range of standard RRT +/- 0.06 units.

3.4.4  Action

Action for retention time evaluation requires professional discretion.  Action taken must be
based on other data such as mass spectra and not on retention times alone. The following
actions are suggested for several potential situations.

      •      Acceptable matching  of  the  mass  spectra  and the  RRTs~No  action
             suggested.
WDCR656/027JI

-------
                                                        Procedure No.:- Ml-SVOA
                                                                      Revision:  1
                                                                 Date:  06/30/1995
                                                                    Page 16 of 19
       •      Acceptable matching of the mass spectra but poor matching of the RRTs-No
              action suggested, but the cause of retention time shift may be investigated.
              Check to,see if similar discrepancies are observed  in other parts  of the
              chromatogram. Often high concentration of a compound will cause shifts in
              the retention times for other compounds in the nearby eluting region,  but
              shifts in retention  times in other areas may not be observed.   Shifts all
              through the chromatogram may be indicative of an erratic system, such as
              flow rate fluctuations, poor temperature regulation, restriction or leakage in
              the system.  . There are likely chances that the sample values in such
              conditions may be inaccurate. Positive data may be qualified as estimated,
              "J", if deemed essential.                   '  ;         I

 3.4.5  Reporting                           ^-:      _•.

 The form presented earlier in Table Ml-SVOA-3 may be used to record any problems in
 the RRT matching.  For the detected compounds in each sample, the calculated RRTs inay
 be recorded under the "RT" column.  A data qualifier code may be added to the  values.
 exceeding 0.06 RRT and requiring qualification, such as  "0.15J", for a compound  with a
 difference of 0.15 RRT units and where a professional.judgment to estimate the data is
'exercised.   •    -..   I        •-,'..      .'.•''..'"'••,•'"'.''    .    •

 3.5 Evaluation of Blanks                                      : :

 Laboratory blanks and field blanks have a profound impact on false positives reported in
 samples; i.e.,  compounds reported as positive detects but not originating from the samples
 themselves. Cross  contamination from the sampling equipment, incidental contamination
 from the field conditions or contamination from the laboratory  equipment or general
 environmental are likely sources of false positives in the samples.
 3.5.1  Acceptance  Criteria

 Criteria for blank evaluation are specified in the EPA's Functional Guidelines.  In addition,
 Region m has some  additional requirements  modifying  the guidance.   The .acceptance
 criteria for blanks apply equally to any type of blanks associated with either sampling or
 analysis, such as trip blanks,  rinsate  blanks,  field or bottle blanks, laboratory  method
 blanks.  While  there are several criteria for evaluating the blanks, the only criterion
 applicable to Level  Ml is the comparison of the blank and sample concentrations.  This
 criterion is as follows:

       For  common  contaminants, such as  the phthalate  esters,  the sample
       concentration must be minimally 10 times the blank concentration to be
       considered a positive detect.  Other contaminants must be present in the

-------
                                                         Procedure No.:  Ml-SVOA
                                                                       Revision:  1
                                                                  Date:  06/30/1995
                                                                     Page 17 of 19
       sample at or above  5 times  the blank concentration before they can be
       considered  significant detects.  The blank with the highest concentration
       must be used (this is, if the laboratory blank has the highest concentration of
       naphthalene, and the equipment rinsate blank has the highest concentration
       of 2-nitroaniline, both of these blanks must be used to qualify the respective
       contaminants).

3.5.2  Review Items

Data requirements and data retrieval  procedures for blanks are the same as those for the
field samples because the blanks as well as field samples are validated similarly.. Form I's,
mass spectra, chjomatograms, quantitationreports, etc.,  are essential for performing a
validation of the blanks first.       >      •• . "* " •.'

3.5.3  Blai& Evaluation  Procedure

Validate the blanks same as the field samples. Detailed validation procedures are described
above under  appropriate'sections. Use the validated blank' results for a comparison with
the sample results.  Make certain that the  samples and blanks are evaluated oh the same
basis of sample .weight or volume, dilution factors, moisture, content, etc.  Use the 5 (or
10) times the highest blank concentrations for qualifying the sample data.

3.5.4  Action

If the sample concentrations do not meet the criteria of 5  (or 10). times the blank
concentration, the  sample results should be considered essentially  undetected (or as not
detected substantially above the levels reported  in the blanks); therefore, flagged "B" in
accordance with  the Region HI data validation guidelines.

3.5.5  Reporting

Form.I's may be used  to write the "B" data qualifier for the data not meeting the blank
criteria.

3.6  Sample Paperwork

The  purpose  for evaluating the sample paperwork is to determine that the samples being
validated are indeed the ones taken  from the  site,  and have  not been tampered with.
Accurate sample identity is of paramount  importance in substantiating the sample data.
Without unequivocal sample identity and chain-of-custody procedures, the sample data may
not be defensible or enforceable.

-------
                                                         Procedure No.: Ml-SVOA
                                                                       Revision:  1
                                                                 Date:  06/30/1995
                                                                     Page 18 of 19
Under the current GLP contracts, the original paperwork (i.e., the purge package or the
administrative record) is .included in the data package from the laboratory.  It is assumed
that the data validator is not privy to the original paperwork; therefore, the evaluation
criteria and procedures described below apply only to the documents that are generally
included in the data validation package.  These documents are the chain-of-custody forms
and Region HI Shipping Record.                            .   •

3.6.1 Acceptance Criteria
                         i  .   ' \  •                       '
Criteria for acceptability or authenticity of the sampling paperwork, document control and
chain-of-custody have been established by the National Enforcement Investigations Center
(NEIQ, in support of the CLP.  Overall criteria are too numerous and subjective to be
discussed  here, but the criteria that apply to data validation are:

       •       The chain-of-custody  form should be properly and completely filleti out
              including the sampler signatures, date and time of sampling, sampling station
              identification, analyses requested, traffic numbers, tag numbers, etc. These
              data are minimally required to confirm the authenticity of the sample and its
              data;   •    ''.          '  •'     '       '."'"''.          '  '.

       •       The chain-of-custody must be maintained at all.times. The custody transfers
              between different parties must be signed and dated.

'3.6.2 Review Items

A copy of the chain-of-custody form originated in the field  and that returned from the
laboratory with the data are essential to confirm the identity of the samples.  In addition,
the Region HI Shipping Record is essential to identify the field QC samples. The chain-of-
custody and Shipping Record are generally located in front of  the data package.

3.6.3 Evaluation Procedure

Ensure that the chain-of-custody form was signed and dated by the samplers, and a time
and date were entered for sample collection.  The laboratory copy of the chain-of-custody
must have the signature of the laboratory sample custodian. Any errors on the form should
have  been crossed out  with a single line through the entry.  Verify that  all collected
samples have unique  station identification,  traffic  numbers  and  sample tag  numbers.
Ensure that the Region :in Shipping Record correctly reflects the information on the chain-
of-custody.

-------
                                                         Procedure No.:  Ml-SVOA
                                                                      Revision:  1
                                                                 Date:  06/30/1995
                                                                     Page 19 of 19
3.6.4 Action
The action to be taken in qualifying the data is highly dependent on the nature of the
problem.  Some errors in paperwork are practically unavoidable in real situations.  An
effort should be made to reconcile the differences by cross checking the field notebooks
against the sampling paperwork.  Occasionally, the samplers may forget to sign the chain-
of-custody;  however,  the field notebooks may  amply  describe  the  sampling event.
Problems are also'inevitable in noting or cross-referencing sample tag numbers and traffic
numbers.  Generally,  there are several alternate sources of information to substantiate or
refute the problem.

3.6.5. Reporting:
                ..      '  '  '      .'  •'        •ป-. *..
Any discrepancies found in the paperwork must be immediately brought to the attention of
the EPA RPM.  Clearly define the problems in a memorandum to the responsible parties.
Attach marked copies of the chain-of-custody forms to substantiate,the findings.

-------
                                                      Procedure No.:  M2-SVOA
                                                                   Revision:  1
                                                              Date:  06/30/1995
                                                                   Page 1 of 39
                                 Appendix B-2
          Validation of Semivolatile Organic Analyte Data
                            Manual Level M2
                        1. Purpose and Applicability

This procedure provides step-by-step instructions to manually Validate the semivolatile
organic analyte (SVOA) data using the manual innovative data validation approach-at Level
M2.. This approach focuses on the use of information contained on the CLP forms and a
review of chromatograms as summarized in Table M2-SVOA-1. The procedures are based
on modifications to Regions Hi's National Functional Guidelines for Organic Data Review.

The procedure is applicable to the SVOA data'pbtaihed using the Contract Laboratory
Program Statement  of Work (CLP SOW). Hard copy data conforming to the 1990 CLP
SOW specifications  are essential in order to cany outthe procedure.

Data validated using  this .procedure  are considered  usable  for  the following types  of
purposes;  however, the  data users must decide on a case-by-case  basis whether the
procedure is suitable, for their intended data uses.  The suggested data uses are:

       •      Oversight of activities led by other parties

       •      Comparison to action levels

       •      Initial site investigation

       •      Contamination sources

       •      Nature and extent of contamination

       •      Preliminary risk assessment

       •      Risk assessment with known high, levels of toxics

       •      Feasibility study

       •      Preliminary design

       •      Treatability study

       •      Initial cleanup verification

-------
                                                               Procedure No.:  M2-SVOA
                                                                             Revision:,  i
                                                                        Date:  06/30/1995
                                                                            Page 2 of 39
                                  Table M2-SVOA-1
                          QC CHECKLIST FOR LEVEL M2
                        CLP RAS SEMIVOLATILE ORGANICS
               QC MEASURES
Action Level Notification
Instrument Tune
Initial Calibration (RRF)
Initial Calibration (96RSD)
Continuing Calibration (RRF)
Continuing Calibration (%D)
Laboratory Blank
MS/MSD (%R, RPD)
Internal'Standard Area
Field Blank
Sample Paperwork
Holding Time
Retention Time
Surrogate Recovery
Dilution Factor
Moisture Content
Mass Spectra
Chromatograms
Raw Data
X

-------
                                                       Procedure No.: M2-SVOA
                                                                    Revision:  1
                                                               Date:  06/30/1995
                                                                   Page 3 of 39

                   2.  Quality Control Measures Checked

Table M2-SVOA-1 highlights the quality control (QC) indicators evaluated under this data
validation procedure.                     '
                                3.  Procedure

The following subsections describe for each of the QC indicators the acceptance criteria,
location and retrieval of QC data, evaluation of the QC data, actions taken in the event the
QC acceptance criteria are exceeded,  and documentation of the  QC violations  ini a
standardised report form.

The semivolatile data requirements to be checked-are listed below:

       3.1    Action Level Notification

       3.2    Technical Holding Times (CCS - Contractual holding times only)

       3.3    GC/MS Instrument Performance Check (CCS)

       3.4    Initial Calibration (CCS)

       3.5    Continuing Calibration (CCS)

       3.6    Blanks (CCS)

       3.7    Surrogate Spikes (CCS)

       3.8    Matrix Spikes/Matrix Spike Duplicates

       3.9    Internal Standards (CCS)

       3.10   Reported Contract Required Quantitation Limits (CRQLs)

       3.11   Tentatively Identified Compounds

Two forms have been  developed  to assist in  the  performance and  documentation of
implementing  Level  M2.  The first form, M2-SVOA-QUAL, summarizes holding time,
calibrations, blanks, surrogates, and internal standards. The second form, M2-SVOA-SPK,
summarizes surrogate and matrix spike quality control checks.  These forms appear on the
following pages as Table M2-SVOA-2 and Table M2-SVOA-3.

-------
                                                           Procedure No.:  M2-SVOA
                                                                         Revision:  1
                                                                    Date:  06/30/1995
                                                                        Page 4 of 39

  Reporting requirements for Level M2 are as follows:
                1      ii
        •      Hand annotate the Form I's, including
                      Data validation qualifiers
                      Sample identification number
                      Sampling location

        •      Provide, a narrative that includes
               -      A statement that defines the level of the data review, i.e., M2
                    ,  Major and minor problems associated with the analysis
                      Highlight issues that may have affected detection limits

        *      Include the following attachments                  v
                      List,of data validation qualifiers  ,       .-.  :'     ,/
               - •     Support documentation including forms that  support assigning data
                    '•qualifiers  ..     \--.•. •   ''  •''     ..  •: ,';_,,'',-,-..;;..  •'._/' -• •':  '  '
                      Chain  of custody form
                      Samples affected fay calibration  should be listed on the appropriate
                    :.  calibration forms                       /'••;';.'.     .   •

 The data qualifiers assigned  in this review are as follows:   .         v,      :

  Codes  Relating  To  Identification  (Confidence concerning presence-or absence of
  compounds)                                   .'•'..    '   •   .   : --

           U  =  Not detected.   The associated number indicates approximate sample
                  concentration necessary to be detected.

(NO CODE)    =  Confirmed identification.

           B  =  Not detected substantially above the level reported in laboratory or field
                  blanks.

           R  =  Unreliable result.  Analyte may or  may not be  present in the sample.
                  Supporting data necessary  to confirm result.

           N  =  Tentative  identification.   Consider  present.  Special methods may be
                  needed to  confirm its presence or absence in future sampling efforts'.

  Codes Related To Quantitation  (can be used for  both positive  results and  sample
  quantitation limits):

           J    =  Analyte present.  Reported value may not be accurate or precise.

-------
                                                Case No.:_
                                                SDG No.:ฃ
                                         Data Reviewer
                                              Site:	
Procedure No.: M2-SVOA

             Revision: 1
        Date:  06/30/1995
               Table M2-SVOA-2. SVGA Qualifier Summary (Calibrations, Blanks, Holding Time, Surrogates, Internal Standards)
Analysts
Date(s):
Analyzed Within
12 Hours of
•nine?
_Yeป__No


IntUlD:


Tune OK?
_i.Yei_No


MBlnkpi
Ext. Dates:
Sample .
Identifier:
l'.
2. ^
3. ' •. .• • " ""
•4;.- •••-•*-$• ;•";<•
5."'" ^ - :'^^-',-
6 ' -•".''• '-'•'•" V
7. - ' • •'"•-•'
g., • ' ' ;.'.:'•
9. '••"'. " -"'''••• :- •
10.' •. •; : ••:•:•„ •-. •
Hold
Time
Out, days
Ex



•..-;-.•.•,-
> •
. , • •
• . ••••

••'!:''
'.. :••-'-•
An



:•-






Standards: (<>>)
Surrogate
1


. '• • .
.--. •. ..
• :•.'' '..'". -
..,.-


' "'

2









. ; .-.
3










4
•









5









• .*"•.'
6







. - .•

"'• '•'•
7



••-.-:
.'"'*• '"^
••.' -'••

•-•- :•

;•_...,. -

8



';.. .
,,•:•!••_' '•
•> '.'.- '

.' ^
- y . '
:". '
"•*"

1





•?t'.-'-


•?•'• •
...i-;. ..
Internal (IS)
2



-'

,;.<-.-,;•



•••• ' . '
3




.,
>:,, •



•- ~
4










5





',



-''•
6










                                                            Time:

Compound; Y SPCC roDane) ' fBN)
4-Methvlohenol fA)
N-Nitroso-di-n-oroovlamine fBN)f#>
Hexachloroethane fBN)
Nitrobenzene ' • fBN)
Isoohorone fBN)
2-Nitroohenol fA)f*V
2.4-Dimethvlohenol fAt
bis(2-Chloroethoxv1methane fBN)
2.4-Dichloroohenol fA)
1.2.4-Trichlorobenzene ' fBN)
Naphthalene fBN)
4-Chloroaniline fBN)
Hexachlorobutadiene fBN)f*)
4-Chloro-3-methvlnhenol - f A)f*)
2-Methvlnaohlhalene fB^f)
HexachlorocvcloDcntadiene fBN)(lf)
2.4.6-Trichloroohenol ' fA)f*)
2.4.5-Trichloroohenol fA)
''-Chloronaohthalene fBN)
2-Nitroaniline fBN)
Dimethvlohthalate " fBN)
Acenaohthvlene fBN)
2.6-Dinitrotoluene fBhf)
3-Nitroaniline fBN)
Acenaohthene f*)fBN)
2.4-Dinitronhenol fA)f#)
4-Nitronhenol fA)(iO
•-.-•>••••:! r,'--'-
RRF30



.































' rnnfiimi
RRF<.05















•



















nztaL" ••
%bvi










"
























v--'vi:v-:-::;
•''-is* -^

• 1






























'


nks •' " '" ' ' • •"'
FSeM




ป.
---~.- •
"•~




























OoaEfiers
'•(+AJ ^








































I
S
#
1







I
S
H
2









I
S
tt
3





-------
                                                    Case No.:_
                                                   SDG No.:	
                                          Data Reviewer:	
                                                  Site:	
                                                  Procedure Noi:  M2-SVOA
                                                               Revision: 1
                                                           Date:  06/30/1995
                                                            Table M2-SVOA-2
                                                                 (contd.)
Analytu Date:
Inftnuneni ID:
                                  Date:
Time:
Compound: ' '•"•.• v^Vr-^SPCC-ttl':
' ..'•:-••.; • ;-:v-^;^^Add!c-(A)T
. B*M/Ntiitnl(BN)
• DibcfBiOflf fl ••*' ' " v' " -' " ""••'• fBN^
2.4-binhrotohieae • '• •' - -';'' •-.'-: -•'•"•YUM) •
Dietiivfohthalite ' •' • •• .~- 7BNV
X-Chloroohenvtohenvleflier (BNV
Fluorene -' • .•- '•••-• •••••'•'"•'"•"•'•'''•'-'• "•(tttn.'
4-Nhraanfline ~ ' \>'->F-"'--'fv:^(Btn!>
4.^ปPinltftv^"ttielhvlphciiol •* -~ *'v".^~- ^-''YA^
N-KhfOioJrohenvlamme ;' v :--' ~~.'< YBN)TT":
4-BromouheBvl-aheiwledier '•' • ' • YBNl

RenttehlomohetMtl '-. • ••' ••''••' •'••(Mft
Phenanihrene '• '•" ' ••• '"."•" (Btfl
-ArttjrปcetiB ' •''•" ' '.'•""• '" ; •""'. -JBr^i
Ca'rbazole • •:''.'.' ':' •• '• ' • IBN>
Dt-n-Butvlnhthalate ' • : ' (Btt\
Fhiofซnthene ' ' • •' (BrDfl
Pvrene " • ' fBN\
_ _. •_ . • •

3 J '-Diehlorab'enndene ' (Btft
BenzoCttanthracene . (Brfl
bij(7-cthvfheTvrtohthal8te flttft
Chrvtene • iBN>
Di-n-Oetvlohthalate (BN1(*1
Benzofblfluoranthene fBNI
BenTortrlfluoranthene • fBH>
Benzofa^Dvrene ' fBKIf*1)
Indenod .2.3-c Jlovrene (BN)
Dibenzfa.h^anthracene • fBN)
Benzofe.h.nDervlene (BN)
^. r; Initial CซI.
>RKr<.05
•' •• . .-. •' ••' .
• -- A ••'.-•• • ' .
'-•^ /'
. . ••• . •

-•v. . • -.-.•• ••;
'.:-.•;•: ':'. '. .'
'.'.'•. •• '. '-. '
•-:.:••• ..' .' "•
'..••'•• ;' '


• -
















%RSD>30






•- .••.••. . -



'-••-. ••





. '



•*








. CoulimnDjE Cu*
RRF<.05

.


- .
~j*- "





















•

ซD>25















.













Blanks • . •
Ub






, 'i
. •: •• " ' :",



• " •

:




.










Field

\ •• •














~-~..,
- 7;- '











Qualifiers
(+/-)'•




(





%



















I
S
I
4
I
S

-------
                                        Case No.:.
                                        SDG No.:_
                            Data Reviewer:	
                                  Site:	
 Procedure No.:  M2-SVOA
              -   Revision:  1
_        Date:  06/30/1995
                     Table M2-SVOA-3. SVGA Soil Surrogate and Matrix Spike Quality Control Summary

                                             Surrogate Data Summary
Sample Identifier:
(Acceptance Ranee, XR):







, • '-•'!. ' ' : .
''..-• • i

Soil Sample Recoveries, %R
SI
23-120








• ...

S2
30-115




"




•-' -
S3
18-137










S4 '
24-113










S5
25-121




' .'


ป• *


S6
19-122










S7
20-130






. •
• '- . •


S8
20-130








: •" -

Qualifiers (+/-)
.. . - .





•
.. •. • , ..


SI - Nhrobenzene-dS;S2 = 2-fluorobiphenyl, S3 •= teiphenyWU. S4 ซ= phenol-d5, S5 = 2-fluorophenol, S6 ซ= 2,4.6-lribromophenol, S7 ซ= 2-
chloropheiiol-d4,S8 - U-dichlotobenzeae-d4,
                                             MS/MSP Pate Summary
SPCC4T)
CCC(ป)
Spike Compound: Aromatic (AR)
SOIL SAMPLES
Phenol
2-Chlorophenol
1 ,4-Dichlorobenzene
N-Nitroso-di-n-propylamine
1 ,2,4-Trichlorobenzene
4-Chloro-3-methylphenol
Acenaphthene
4-Nitrophenol
2.4-Dtnitrotoluene
Penlachlorophenol
Pyrene
Matrix Spike, Recovery.
%R
Range

26-90
25-102
28-104
41-126
38-107
26-103
3M37
11-114
28-89
17-109
35-142
Actual












Matrix Spike Duplicate,
Recovery* %R
Range

26-90
25-102
. 28-104
41-126
38-107
26-103
31-137
11-114
28-89
17-109
35-142
Actual












MS/MSD
Precision, RPD
Range

35
50
27 .
38
23 .
33
19
50
47
47
36
Actual












Qualifiers
(+/-)













-------
                                         Case No.:.
                                       .  SDG No.:.
                             Data Reviewer:	
                                   Site:	
 Procedure No.:  M2-SVOA
                  Revision:  1
_         Date:  06/30/1995
                    Table M2-SVOA-3. SVGA Aqueous Surrogate and Matrix Spike Quality Control Summary

                                             ^Surrogate Data Sfimmarv
Sample Identifier:
(Acceptance Ranee. XR):
1. . ' .•- '••
2. "' ' . ' '•
3. .. . "•
4. '
5.
6.
7.
s. ' •: . ..
9."'" ' .'-'•
10.
Aquebui Sample Recoveries
Si
35-114










S2
43-116







.' • -,:-

•--.'
S3
33-141




-
• •



.<;• .
S4
10-110








,

S5
21-110






;-.':
ป• *" .


*R
S6
10-123





. - ' '.




S7
33-110

i

•i






S8
16-110
•-. • ' ..









Qualifier* (+/-)

\



^ ^ '

-..'.. ;
-".""'•• ' •' -

SI •> Nhrobenzene-
-------
                                                        Procedure No.: M2-SVOA
                                                                      Revision:  1
                                                                Date:  06/30/1995
                                                                     Page 9 of 39

        K   =  Analyte present./Reported value may be biased high.  Actual value is
                 expected lower.

        L   =  Analyte present.  Reported-value may be biased low.  Actual value is
                 expected to be higher.      '   .  '   ;  .

       UJ   =  Not detected, quantitation limit may be inaccurate or imprecise.

       UL   =  Not detected, quantitation limit is probably higher.

Other Codes

        Q   =  No analytical result   .

         *   =  Results reported from diluted analysis.
                             '   "         '  .    •'
3.1  Action Level Notification

The purpose behind action level,notification is to make the EPA Remedial Project Officer
(RPM) or the Site Project Officer (SPO) aware of the potential human health risk  at the
site.  In accordance with the Region HI Hazardous waste division policy, the EPA RPM or
SPO must be promptly notified of any contaminant exceeding the established action level or
the 10-day health advisory limit. The data  for contaminants exceeding tte action  levels
must be validated as a top priority and reported to the RPM or SPO,  Validation of the rest
of the data may then be completed normally.

3.1.1  Acceptance Criteria

EPA's Office of Solid Waste and Emergency Response has established 10-day advisory
limits or the action levels for several organic compounds and elements of special health risk
concerns based on the Safe Drinking Water Act.  The semivolatile organic compounds and
their 10-day health advisory limits apply only to aqueous samples and are listed in  Table
M2-SVOA-4. The criteria for action level notification are as follows:

       •      The contaminant concentration must be equal to or above the established 10-
             day health advisory limits.

       •      Data for contaminants exceeding the action levels must be validated as a top
             priority.

       •      The following  EPA  personnel  must   be notified  of  the  action  level
             exceedances:

-------
                                          Procedure No.:  M2-SVGA
                                                       Revision:  1
                                                  Date:  06/30/1995
                                                     Page 10 of 39

      EPA RPM or SPO

      EPA Section Chiefs:

      -     Site Investigation (SI)

      r-  .   Remedial

      :*-; ; ;  Enforcement

      -T     RCRA

      EPA Section Toxicologists:

      T-  .v  Enforcement

      —     Superfund

      r     RCRA

The remaining data validation should be completed per normal procedures.

Any special instructions  from the  Hazardous  Waste Division should be
followed.           .

Records should be kept of the data review, action level notification and any
follow up instructions and actions.
Table M2-SVOA-4
SEMIVOLATILE ORGANIC ANALYTES AND ACTION LEVELS
Compound
1 ,3-Dichlorobenzene
Pentachlorophenol
Action
Level*
8,930
300
Compound
. 1,4-Dicnlorobenzene
—
Action
Level*
10,700
—
*A11 units are ug/1.

-------
                                                        Procedure No.:  M2-SVOA
                                                                      Revision:  1
                                                                 Date:  06/30/1995
                                                                     Page 11 of 39
3.1.2  Review Items
All data required to perform Level M2 validation, as detailed in the following sections, are
necessary for carrying out action level notification.  The location of. the data and their
retrieval procedures are also discussed below.

3.1.3 Evaluation  Procedure

The evaluation process preceding action level  notification will  primarily  consist of
comparing the results on Form I's with the action levels presented in Table M2-SVOA-4.
Following the identification of the contaminants exceeding the action levels, focused data
validation should  be performed  using  the  criteria and procedures described in  the
appropriate sections below.

3.1.4 Action

The action resulting  from  focused data validation will be the notification of action level
exceedance to the personnel identified above in Section 3.1.1.

3.1.5 Reporting

Copies of Form I's can be used to highlight the contaminants above the action levels. The
findings of the focused validation can be summarized  in a memorandum," and  the data
qualifiers resulting  from focused validation may be written on the Form I's.  The marked
up  forms should  be clarified that they represent validation of only the  contaminants
exceeding the action levels, and not all data;
3.2  Technical Holding Times

The objective is to ascertain the validity of results based on the holding time of the sample
from time of collection to time of sample extraction and analysis.

3.2.1  Acceptance Criteria

Technical requirements for sample holding  times have only been established for water
matrices.  The holding times for soils (and "other non-aqueous matrices such as sediments,
oily wastes,  and  sludge) are currently under  investigation.  When the results are available
they will be  incorporated into the data evaluation process.  Additionally, results of holding
time studies  will be incorporated into the data review criteria as the studies are conducted
and approved.

-------
                                                       Procedure No.: M2-SVOA
                                                                     Revision:  1
                                                                Date:  06/30/1995
                                                                   Page 12 of 39

The holding time criteria for water samples, as stated in the current 40 CFR Part 136
(Clean Water Act) is as follows:

       For semivolatile compounds in cooled (@ 4ฐQ water samples the maximum
       holding time is 7 days from sample collection to extraction and 40 days from
       sample extraction to analysis.

It is further  required  that  semivolatile compounds k properly preserved  non-aqueous
samples be extracted within 7 days from sample collection and the .extracts analyzed within
40 days from sample extraction.

3.2.2  Review Items

Form I SV-1 and SV-2, EPA Sample Shipping Log -and/or chain-of-custody.

3.2.3  Evaluation Procedures

Technical holding times for sample extraction are established by comparing the sampling -
date on the EPA Sample Traffic Report with the dates of extraction on Form I SV-1 and/
SV-2.  To determine if the samples were analyzed within the holding time after extraction,
compare the dates of extraction on the sample extraction sheets with the dates of analysis
on Form I SV-1 and SV-2.

Verify  that the C.O.C. indicates that the samples  were received intact and iced. If the
samples were not iced or there were any problems with the samples  upon receipt, then
discrepancies in the sample condition could affect the data.

3.2.4  Action
1.    If technical Holding times are exceeded, flag all positive results as estimated "J"
      and sample quantitation limits as estimated "UJ" and document that holding times
      were  exceeded.   However, please  note  that some extractable  compounds are
      extremely persistent in the environment (e.g., PAHs) in non-aqueous matrices and
      would not be expected to- degrade significantly during sample  storage.   The
      reviewer must use professional judgment  in the application of data qualifiers to
      those compounds in non-aqueous matrices.

2.    If in the professional judgment  of the  data  reviewer  a loss  of semivolatile
      compound(s)  is evident due to exceeding the holding time criteria, the affected
      positive results or the associated quantitation limits may be qualified as biased low,
      "L" or "UL" respectively. The narrative must contain the reviewer's justification
    -  for qualification of the compound results as biased low.

-------
                                                        Procedure No.:- M2-SVOA
                                                                      Revision:  1
                                                                 Date:  06/30/1995
                                                                    Page 13 of 39.

3.     If technical holding times are grossly exceeded (greater than 2 times the required
       technical holding time), either on the first analysis or upon re-analysis, the reviewer
       must use professional judgment to determine the reliability of the data and the
       effects of additional storage on the sample results. The reviewer may determine
       that positive results or the associated quantitation limits are approximates and should
       be qualified with "J" or "UJ", respectively.  The reviewer may determine that
       non-detect data are unusable (R). ,

4.     Because of limited information concerning holding times for non-aqueous samples,
       it is recommended, that a comment in the data review narrative be included to state
       that aqueous holding times were applied.

5.     Whenever possible, the reviewer should comment on the effect of exceeding the
       holding time oh the resulting data in the data review narrative.
'      '•'.-"   V'   ''*••;'      " -.'.                       - "
6.     When contractual and/or technical holding times are exceeded, this should be noted
       on me ORDAS form. 7

7.     The reviewer should also be aware of the scenario in  which the laboratory has
       exceeded the technical holding times, but met contractual holding times.  In this
       case,  the data .reviewer  should notify  the Regional TPO (where samples were
       collected) and/or RSCG that shipment delays may have  occurred so that the field
       problem can be corrected.  The reviewer may pass this information on to the
       Regional TPO on the ORDAs, but should explain that contractually the laboratory
       met the requirements.

8.     When there  are other quality control problems in conjunction with exceeded holding
       times (such  as suspected laboratory contamination), the reviewer should follow the
       hierarchy of qualifiers.   In particular, if for any reason the reviewer doubts the
       presence of a compound, the data summary should display only the "B"  or "R"
       qualifier, and not the "L" qualifier.  This is because no net direction of bias can be
       inferred under these conditions.

3.3  GC/MS Instrument Performance Check

Gjas chromatograph/mass spectrometer (GC/MS) instrument performance checks (formerly
referred to as tuning) are performed to ensure mass resolution, identification and, to some
degree, sensitivity.   These criteria are not sample specific. Conformance is determined
using standard materials, therefore, these criteria should be met in all circumstances.

-------
                                                     Procedure No.:  M2-SVOA
                                                                  Revision:  1
                                                             Date: 06/30/1995
                                                               .  Page 14 of 39
3.3.1  Acceptance Criteria
The analysis of the instrument performance check solution must be performed  at the
beginning of each  12-hour period during which samples or standards are analyzed. The
instrument performance check,  decafluorotriphenylphosphine (DFTPP) for semivolatile
analysis, must meet the ion abundance criteria given below.    ,

            Decafluorotriphenylphosphine (DFTPP)

            m/z          ION ABUNDANCE CRITERIA

            51           30.0-80.0% of m/z 198
            68           Less than 2.0% of m/z 69
            69  '         Present           *•".':/
            70           Less than 2.0% of m/z 69
             127          25.0-75.0% of m/z 198
             197          Less than 1JO% of m/z 198 ,
            198          Base peak, 100% relative abundance
             199          5.0-9.0% of m/z 198
            275          10.0-30.0% of m/z 198
            365          Greater than 0.75% of m/z 198
            441          Present, but less than m/z 443
            442          40.0 -110.0% of m/z 198
            443          15.0-24.0% of m/z 442

NOTE:      All ion abundances must be normalized to m/z 198, the nominal base peak,
            even though the ion abundances of m/z 442 may be up to 110 percent that of
            m/z 198.     •

3.3.2  Review Items

FormVSV.

3.3.3  Evaluation Procedures

1.    Compare the data presented on each GC/MS Instrument Performance Check (Form
      V SV) with  each mass listing submitted and ensure the following:

      a.     Form V SV is present and completed fgr each. 12-hour period during  which
            samples were analyzed.

-------
                                                        Procedure No.:  M2-SVOA
                                                                      Revision:  1
                                                                 Date: 06/30/1995
                                                                     Page 15 of 39

       b.    The appropriate number of significant figures has been reported (number of
             significant figures given for each ion in the ion abundance criteria column)
             and that rounding is correct.

       c.    The laboratory has not made any calculation errors.
             '           *"'".'     '•        '       "          .'•''-'.
2.     Verify that the ion abundance criteria were met. The criteria for m/z 68, 70, 441,
       and 443 are calculated by normalizing to the specified m/z.
 ."  • •'„"••  ? • • > '  •'', •,'•.'"' •"' - •• • •""" r":••'_ 'V1.',' :•""   *••'.'.         '   ;   •     •   •     • •
      . ' -      •"••"•'.'•'.."••_-   • •    -      .          \  .    i"   .'•'••
3.     If possible,  verify  that spectra were generated using appropriate background
       subtraction  techniques.     Since   the DFTPP  spectrum   is  obtained  from
       chromatographic peaks that should be  free from coelutioh problems, background
 -     subtraction should be done in accordance with the following procedure;   Three
       scans (the peak apex scan and the scans .immediately preceding and following the
       apex) are acquired and averaged and background subtraction must be accomplished
       using a single/scan prior to the elution of DFTPP.       •       v    .    ?

       NOTE: '     All  instrument conditions must  be identical to those used  in  the
                ?   sample analysis.  Background subtraction actions resulting in spectral
                    distortions for the sole purpose of meeting the contract specifications
                    are contrary to  the quality assurance objectives and are therefore
       •             unacceptable.

3.3.4 Action

1.     If the laboratory  has made minor transcription errors which do not significantly
       affect the data, the data reviewer should make the  necessary corrections on a copy
       of the form.                                              .

2.     If the laboratory has failed to provide  the correct forms or has made significant
       transcription  or calculation errors,  the  Region's designated representative should
       contact  the laboratory and request corrected  data.   If  the  information  is  not
       available, then the reviewer must use professional judgment to assess the data. The
       Regional TPO should be notified by noting the problem(s) on the ORDAS.

3.     If mass assignment is in error (such as m/z 199 is indicated as the base peak rather
       than m/z 198), classify all associated data as unusable, (R).

4.     If ion abundance criteria are not met,  professional judgment  may  be  applied to
       determine  to  what extent  the data  may be utilized.   Guidelines  to aid in  the
       application of professional judgment in evaluating ion  abundance criteria  are
       discussed as follows:

-------
                                                        Procedure No.:  M2-SVOA
                                                                      Revision: 1
                                                                 Date:  06/30/1995
                                                                     Page 16 of 39

       a.     Some  of the  most  critical  factors in  the DFTPP  criteria  are  the
             non-instrument specific requirements that are also not unduly affected by the
             location of the spectrum on the chromatographic profile.  The m/z ratios for
             198/199 and 442/443 are critical.  These ratios are based on the natural
             abundances of  carbon 12  and carbon  13 and, should always  be met.
             Similarly, the relative abundances for m/z 68, 70,197, and 441 indicate the
             condition of the instrument and the suitability of the resolution adjustment
             and are very important Note that all of the foregoing abundances relate to
             adjacent ions;  they are relatively insensitive to differences in instrument
             design: and position of the spectrum on the chromatographic profile. .

       b.     For the ions at m/z 51,127, and 275, the actual relative abundance is not as
             critical.  For instance, if m/z 275 has 40% relative abundance  (criteria:
             10.0-30.0%) and other criteria arejnet, then the deficiency is minor.

       c.    .The relative abundance of m/z 365 is an indicator of suitable instrument zero
             adjustment.  If relative abundance  for m/z 365 is zero, minimum detection
             limits may be affected. On the other hand,  if m/z 365 is present, but less
             than the 0.75% minimum  abundance criteria,  the deficiency is not as-
      •      .serious/    .•  .     '••.    • •     .      • . •••' -.'-      '.':.'    •  ':

5.     Decisions to use analytical data associated with DFTPP instrument performance
       checks not meeting contract requirements should be clearly noted in the data review
       narrative.    \  .                      ,                       ' ~-

6.     If the reviewer has reason to believe that instrument performance  check criteria
       were achieved using techniques  other than  those  specified in  the SOW and in
       subparagraph  a.  above,   additional  information  on  the  DFTPP  instrument
       performance checks should be obtained. If the techniques employed are found to be
       at variance with contract requirements, the procedures of the laboratory may merit
       evaluation.   Concerns  or  questions regarding laboratory performance should be
       noted for TPO action on the ORDAS.   For example, if the reviewer has reason to
       believe that an inappropriate technique was used to obtain background subtraction
       (such as background subtracting from  the solvent front or from another region of
       the chromatogram rather than the DFTPP peak), then this should be noted for TPO
       action on the ORDAS.                                   ,

3.4  Initial Calibration

Compliance requirements for satisfactory instrument calibration are established to ensure
that the instrument is capable of producing acceptable qualitative and quantitative data for
compounds  on the  semivolatile  Target  Compound List (TCL).    Initial  calibration

-------
                                                         Procedure No.:  M2-SVOA
                                                                      Revision:  1
                                                                 Date:  06/30/1995
                                                                     Page 17 of 39

demonstrates that the instrument is-capable of acceptable performance in the beginning of
the analytical run and of producing a linear calibration curve.

3.4.1 Acceptance Criteria           • .;.  •

1.     Initial calibration  standards containing  both  semivolatile  target compounds  and
       surrogates are analyzed at concentrations of 20, SO, 80,  120, and 160  ug/L at the
       beginning of each analytical sequence or as necessary if the continuing calibration
       acceptance criteria are not met. The initial calibration (and any associated samples
       and blanks) must  be analyzed within 12 hours of the  associated  instrument
       performance check.

2.     Minimum Relative Response Factor (RRF) criteria must be greater than or equal to
       0.05.  Contractual RRF criteria are Iisted4ir Appendix A.

3.     The Percent Relative Standard Deviations  (%RSD) for the RRFs  in the initial
       calibration must be less than or equal to 30%.

3.4.2 Review Items

Form VI SV-1 and SV-2 and chromatograms.

3.4.3 Evaluation Procedures

1.     Verify that the  correct concentration  of  standards  were used  for  the initial
       calibration (i.e., 20, 50,. 80, 120, and 160 ug/L)*  For the eight compounds with
       higher CRQLs, only a four-point initial  calibration is required (i.e., 50,  80, 120,
       and 160 ug/L).                                        '

2.     If any sample results were calculated using an initial calibration, verify that the
       correct standard (i.e., the 50 ppb standard) was used for calculating sample results
       and that the samples were analyzed within  12 hours of  the associated instrument
       performance check.

3.     Evaluate the RRFs for all semivolatile target compounds and surrogates:

             Verify that all semivolatile target compounds and surrogates have RRFs that
             are greater than or equal to 0.05.  If problems are  suspected with low
             response factor or compound identification, also check elution  order.

       NOTE:      Because historical performance data indicate poor response  and/or
                   erratic behavior, the semivolatile compounds listed above have no

                                                          U.S. EPA Headquarters Library
                                                                 Mail code 3201
                                                          1200 Pennsylvania Avenue NW
                                                             Washington DC  20460

-------
                                                        Procedure No.:  M2-SVOA
                                                                     Revision:  1
                                                                Date: 06/30/1995
                                                                    Page 18 of 39

                   contractual maximum %RSD criteria.  Contractually they must meet
                   a minimum  RRF criteria of  0.01;  however, for  data review
                   purposes, the "greater than or equal to 0.05" criterion, is applied
                   to aU semivolatile compounds.

              Semivolatile Target Compounds Exhibiting Poor Response
               '     : '      '     '      '          .   ' '    •
                   i^'-oxybis^-Chlorppropane)
                  , 4-Chloroaniline
                   Hexachlorobutadiene  ,
                   Hexachlorocyclopentadiene
                   2-Nitroaniline
                   Dimethylphthalate
                   3-Nitroaniline
                   2,4-Dinitrophenol
                   4-Nitrophenol
                   Carbazole
                   Diethylphthalate
             •     4-Nitroaniline         ^
             •     4,6-Dinitro-2-methylphenol
             •     N-Nitrosodiphenylamine
             •     Di-n-butylphthalate
             •     Butylbenzylphthalate
             •     3-3'-Dichlorobenzidine
             •     bis(2-Ethylhexyl)phthalate
             •     Di-n-octylphthalate

4.    Evaluate the %RSD for all semivolatile target compounds and surrogates.

      a.     Verify that all semivolatile target compounds have a %RSD of less than or
             equal to 30%.  The contractual criteria for an acceptable initial calibration
             specifies that up to any 4 semivolatile  target compounds may fail to meet
             minimum RRF or maximum %RSD as long as they have RRFs  that are
             greater than or equal to 0.010, and %RSD of less than or equal to 40.0%.
             For data review purposes, however, all compounds must be considered for
             qualification when the %RSD exceeds the ฑ  30.0% criterion.

      b.     If  the  %RSD is  greater than 30.0%, then  the reviewer should  use
             professional judgment to determine the need to check the points on the curve
             for the cause of the non-linearity. This is checked by eliminating either the
             high point or the low point and recalculating the %RSD.

-------
                                                         Procedure No.:  M2-SVOA
                                                                       Revision:  1
                                                                  Date:  06/30/1995
                                                                   .   Page 19 of 39
3.4.4 Action
1.     All semivolatile target compounds, including  the 19 "poor performers," listed
       above, will be qualified using the following criteria:

       a.     If the %RSD is greater than 30.0% and the RRF is greater than or equal to
             0.05, qualify positive results with "J", and non-detected semivolatile target
             compounds using professional judgment

       b.     If the'RRF  is less than 0.05, qualify positive results that have acceptable
             mass spectral identification with "J" using professional judgment, and non-
             detects as unusable.(R).   '.

2.     At the reviewer's discretion, a more in-depth review to minimize the qualification
       of data can be accomplished by considering the following:

       a.     If any of the required semivolatile compounds have a %RSD greater than
             30.0%, and  if eliminating either the high or the low point of the curve does,
             hot restore the %RSD to less than or equal to 30.0%:  .

             i.      Qualify positive results for that compound(s) with "J".

             ii.     Qualify  non-detected semivolatile  target  compounds  based on
                   professional judgment.                        •/--'-

       b.     If the high point of the curve is outside of the linearity criteria (e.g. due to
             saturation):

             i.     No qualifiers are required for positive results in the linear portion of
                   the curve.

             ii.    Qualify positive results outside of the linear portion of the curve with
                    ซซTป>                 •          .   '
                     J .

             iii.   No qualifiers are needed for non-detected target compounds.

       c.     If the low end of the curve is outside of the linearity criteria:

             i.     No qualifiers are required for positive results in the linear portion of
                   the curve.
                                                   t

             ii.    Qualify low level positive results in the area of non-linearity  with
                   "J".

-------
                                                         rroceuurc i>o.:  M2-SVOA
                                                                      Revision:  1
                                                                 Date:  06/30/1995
                                                                     Page 20 of 39

             ill.    Qualify  non-detected   semivolatile   target   compounds   using
                    professional judgment.

3.     If  the  laboratory has  failed to provide adequate  calibration information,  the
       designated representative  should contact the laboratory and request the necessary
       information. If the information is not available, the reviewer must use professional
       judgment to assess the data.

4.     Whenever possible, the potential effects on the data resulting from a failure to meet
       calibration criteria should  be noted in the data review narrative.

5. •    If calibration criteria are grossly exceeded, this should be noted for TPO action on
       the ORDAS.

'6.     When it is suspected that relative response factors were incorrectly generated from
       misidentified peaks  or  incorrect area measurements,  the laboratory  should be
       contacted to requantitate these RRFs and associated sample results. .The ORDAS
       .should identify affected results and document the cause of the reviewer's suspicions.
       In addition, a CLP telephone log must be completed.

7.     Positive results for compounds flagged for blank contamination (B) will not need a
       separate flag (J)  in the data summary form for minimum RRF, %RSD, or %D
       outside criteria.  However, these situations should be addressed UL the data review
       narrative and issues pertaining  to noncompliance should be  documented  on the
       ORDAS.

3.5  Continuing Calibration

Compliance requirements for satisfactory instrument calibration are established to ensure
that the instrument is capable of producing  acceptable  qualitative and quantitative data for
semivolatile target compounds.  Continuing .calibration  establishes  the 12-hour relative
response factors on which the quantisations  are based and checks satisfactory performance
of the instrument on a day-to-day basis.

3.5.1  Acceptance Criteria

1.     Continuing  calibration standards  containing both target compounds and surrogates
       are analyzed at the beginning of each 12-hour analysis period following the analysis
       of  the  instrument performance  check and prior to the analysis of  blanks  and
       samples.

2.     The minimum Relative Response Factors  (RRF) for semivolatile target  compounds
       and surrogates must be greater than or equal to  0.05.

-------
                                                       Procedure No.: M2-SVOA
                                                                     Revision:  1
                                                                Date:  06/30/1995
                                                                   Page 21 of 39

3.     The percent difference (%D) between the initial calibration RRF and the continuing
       calibration RRF must be within +.25.0% for all target compounds.

3.5.2 Review Items

Form VH SV-1 and SV-2 and chromatograms.

3.5.3 Evaluation Procedures

1.     Verify that the continuing calibration was run at the required frequency and that the
       continuing calibration was compared to the correct initial calibration.

2.     Evaluate the continuing calibration RRF for all semivolatile target compounds and
       surrogates.     .                     •*"'•'•            :

             Verify that all semivolatile target compounds and surrogates have RRFs
             within specifications.

    NOTE:  Because .historical performance data indicate poor response and/or erratic
             behavior,  the compounds  listed  in  Section  3.4.3 have no  contractual
             maximum  %D criteria.  Contractually they must meet a minimum RRF
             criterion of 0.01; however, for data review purposes, the "greater than
             or equal to 0.05" criterion is applied to all semivolatile: compounds.

3.     Evaluate the %D between initial calibration RRF  and continuing calibration RRF
       for one or more semivolatile compounds.

             Verify that the %D is within the +_  25.0% criterion, for all semivolatile
             target compounds and surrogates.  Note those compounds which have a %D
             outside the +.25.0% criterion.  The  contractual criteria for an acceptable
             continuing  calibration  specifies  that  up  to  any 4  semivolatile target
             compounds may fail to meet minimum RRF or maximum %D as long as
             they have RRFs that are greater than or equal to 0.010, and  %D of less than
             or equal to 40.0%.  For data review purposes, however, all compounds must
             be considered for qualification  when the %D  exceeds  the  ฑ25.0%
             criterion.

3.5.4 Action

1.     The reviewer should use professional judgment to determine if it is necessary  to
       qualify  the data for any  semivolatile target compound.  If qualification of data is
       required, it should be performed using the following guidelines:

-------
a.
                                                         Procedure No.: M2-SVOA
                                                                       Revision:  1
                                                                 Date:  06/30/1995
                                                                     Page 22 of 39

             If the  %D is outside the ฑ 25.0% criterion and the continuing calibration
             RRF is greater than or equal to 0.05, qualify positive results "J".

       b.     If the  %D is outside the ฑ 25,0% criterion and the continuing calibration
             RRF is greater than or equal to 0.05, qualify non-detected semivolatile target
             compounds based on professional judgment      .

       c.     If the continuing calibration RRF is less than 0.05,  qualify positive  results
          .   that have acceptable  mass  spectral identification  with  "J"  or  use
             professional judgment.,          -

       d.     If the  continuing calibration RRF is less than 0.05, qualify non-detected
             semivolatile target compounds as unusable (R).

2.     If  the  laboratory has  failed to provide adequate calibration  information, the
       designated representative should contact  the laboratory and request the necessary
       information. If the information is not available, the reviewer must use professional
       judgment to assess'the data.

3.     Whenever possible, the potential effects on the data resulting from a failure to meet
       calibration criteria should be noted in the data review narrative.

4.     If calibration criteria are grossly exceeded, this should be noted for TPO action on
       theORDAS.                                                  c

5.     When it is suspected that relative response factors were incorrectly generated from
       misidentified  peaks  or incorrect area measurements, the laboratory should be
       contacted to requantitate these RRFs and associated sample results.  The ORDAS
       should identify, affected results and document the cause of the reviewer's suspicions.
       In addition, a CLP telephone log must be completed.

6.     Positive results for compounds flagged for blank contamination (B) will not  need a
       separate flag  (J)  in the data summary form  for minimum  RRF, %RSD, or  %D
       outside criteria.  However, these situations should be addressed in the data review
       narrative and issues  pertaining  to noncompliance should  be documented  on the
       ORDAS.

3.6  Blanks

The  purpose of laboratory (or field) blank  analyses is to determine the existence and
magnitude of contamination problems resulting from laboratory (or field) activities.  The
criteria for evaluation of blanks apply to any blank associated with the samples (e.g.,

-------
                                                        Procedure No.: M2-SVOA
                                                                      Revision:  1
                                                                 Date:  06/30/1995
                                                                    Page 23 of 39

method blanks, instrument blanks, trip blanks, and equipment blanks). If problems with
any blank exist, all associated data must be carefully evaluated to determine whether or not
there is an inherent variability in the data, or if the problem is an isolated occurrence not
affecting  other data.

3.6.1  Acceptance Criteria

1.     No contaminants should be found in the blanks.

2.     The method blank must be analyzed on each GC/MS system used to analyze that
       specific group or set of samples.

3.6.2  Review Items
       '       '   '   ..  •        .     . -   '  •*• " •  .  • '
Form I SV-1 and SV-2, Form IV SV and chromatograms.
                     ,    .   -,  •    .            .
3.6.3  Evaluation  Procedures

1.     Review  the results  of all  associated  blanks,  Form  I  SV-1  and  SV-2, and
       chromatograms to evaluate the presence of target and non-target compounds in the
       blanks.

2.     Verify that a method blank analysis has been reported per matrix, per concentration
       level, for each extraction batch and for each GC/MS system used to analyze
       semivolatile samples. The reviewer can use the Method Blank Summary (Form IV
       SV) to assist in identifying samples associated with each method blank.

3.6.4  Action

If the appropriate blanks were not analyzed with the frequency  described above, then the
data reviewer should  use professional judgment to determine if the associated sample data
should  be qualified.  The reviewer may  need to obtain  additional information  from  the
laboratory.  The situation should be noted for TPO action on the ORDAS.

Action  in the case of unsuitable blank results depends on the circumstances and  origin of
the blank.  Positive  sample results should be reported  unless the concentration of the
compound in the sample is less than or equal to 10 times (lOx) the amount in any  blank for
the common phthalate contaminants, or 5 times the amount for  other compounds.   In
instances  where more than one blank is associated with a given sample, qualification should
be based  upon a comparison with the associated blank* having the highest concentration of
a contaminant.  The results must not be corrected by subtracting any blank value.

-------
                                                         Procedure No.:  M2-SVOA
                                                                       Revision:, 1
                                                                 Date:  06/30/1995
                                                                     Page 24 of 39

             For qualification purposes, to determine the highest  concentration of  a
             contaminant, consider 'all blanks in a case associated with all samples.

Field blanks measure contamination introduced not only in the  field but also from  the
laboratory.  In  general, evaluation of the impact on specific sample results, is handled as
with laboratory blanks. The reviewer should use caution in attributing contamination to the
field as opposed to laboratory sources.  However,  when field-introduced contamination is
suspected, it is  helpful for the reviewer to consult the sampling group to identify possible
sources and prevent future reoccurrences.  Verified field sources of contamination  should
be noted in the data review narrative.  If a  field blank has a highest concentration of a
contaminant, then'all samples in the associated case are qualified "B",  using the Sx and
lOx rule.  Other field blanks associated with the case are not qualified.

Specific actions are as follows:

1.     If a semivolatile compound is found in a blank but not found in the sample,  no
       action is taken.  If the contaminants  found are volatile target compounds  (or
       interfering non-target compounds) at significant concentrations above  the CRQL,
       then this should be noted for TPO action on the ORDAS.

2.     Any  semivolatile compound  detected in the  sample (other  than  the common
       phthalate contaminants), that was also  detected in any associated blank, is qualified
       "B" if the sample concentration is less than five times (Sx)  the blank concentration.
       For phthalate contaminants, the results are qualified "B" when  the sample result is
       less than lOx the blank concentration.

       In using the 5x/10x rule to compare blank  results  to sample  results which were
       calculated  using different weights, volumes, or dilution factors, the reviewer must
       choose between comparing the levels detected with the instrument, the total amount
       of compound (ug of contamination) present in the extracts, or the final concentration
       of the contaminant in the sample aliquots. Often, more than one  approach will be
       acceptable and will yield the equivalent flagging of sample results.

       a.     Comparisons involving sample dry weight correction factors, but with  all
             other calculation factors the same for sample versus  blank:

                    In this case,  the reviewer can compare the wet weight
                    concentrations, instrument levels, or the total amount
                    of compound (ug of contaminant) in the extracts. All
                    of these approaches will be acceptable and will  yield
                    equivalent flagging of sample results.

       b.     When the sample has a smaller initial  aliquot size than the blank (purge or
             extraction weight/volume), but all other  calculation factors  beyond this

-------
                                                  Procedure No.:  M2-SVOA
                                                                Revision:  1
                                                            Date: 06/30/1995
                                                               Page 25 of 39

       analytical  step  are .identical (Le., same final extract volumes,  injection
       volumes, and extract dilution factors for sample versus blank):

       •     In  this  case, it  is  acceptable and  equivalent to compare  either
             instrument levels, the total amount of compound (ug of contaminant)
             in the extracts, or the concentration of contaminant in the extracts.

       • '••';.   Final concentrations of sample versus blank should not be compared.

c.     When the sample has a larger final extract volume or a greater  dilution
       factor than the blank:        . '       >;          •        .

     '•::*(--^   If the laboratory contaminant;may have been  introduced after or
             during the  sample  dilution-' step,  then  a  direct comparison  of
      ,       instrument levels  is  appropriate.   OFor 'example, comparing  the
      :       instrument level result for a water sample that was diluted  1:100
             prior to injection  would'_ take  into account possible  laboratory
             contamination of the syringe, instrument, or dilution solvent

       •     On the other hand, if it is highly probable  that the contamination
             originated before the dilution step, then it is more  appropriate to
             calculate and compare the  total amount  of  compound  (ug  of
             contaminant) present in the undiluted extract of the sample versus the
             blank. For example, a SNA extract diluted 1:100 prior to injection
             may only be subject to phthalate contamination prior  to the  dilution
             step (i.e., during extraction/concentration).  •

       •     If the results  of a dilution run are to be flagged (B) because of blank
             contamination, the reviewer should attempt to determine whether an
             undiluted run was also performed.  If so, the undiluted run  may be
             used to verify the presence of a compound detected at levels too high
             to be  questioned or,  conversely, to prove  that a compound was
             actually not present at levels multiplied by a dilution factor.

The reviewer should note that blanks may hot involve the same weights, volumes,
or dilution factors as the associated samples.  These factors  must  be taken into
consideration when applying the "5x" and "10x" criteria, such that a comparison
of the total amount of contamination is actually made.

Additionally, there may be instances where little or no contamination  was present in
the associated blanks, but qualification of the sample was deemed necessary.
Contamination introduced  through  dilution is one example.   Although it is  not
always possible to determine,  instances  of this occurring  can be detected  when
contaminants are found in the diluted sample result,  but are  absent in the undiluted

-------
                                                         Procedure No.: M2-SVOA
                                                                       Revision:  1
                                                                 Date:  06/30/1995
                                                                     Page 26 of 39

       sample result.  Since both results are not routinely reported, it may be impossible to
       verify this source of contamination.  However, if the reviewer determines that the
       contamination is from a source other than the sample, he/she  should qualify the
       data. An explanation of the rationale used for this determination should be provided
       in the narrative accompanying the Regional Data  Assessment Summary.

3.     If gross  contamination  exists (i.e.,  saturated  peaks  by GC/MS),  all  affected
       compounds in  the associated samples should be  qualified as  unusable (R), due to
       interference. This should be noted for TPO action if the contamination is suspected
       of having an effect on the sample results.            .-'.'••;.  . ( .  , •

4.     If inordinate amounts of other target compounds are found at  low levels in the
       blank(s), it may be indicative of a problem and should be noted for TPO action.
       '.••'"  '•:-'-".^-'"  ^>;:T'^v^^.r-:jW^-4'^^^:-:.'^';3:^-";'-'''•••' '   '   .   •'
5.     The same consideration given to the target compounds should .also  be given to
       Tentatively Identified Compounds (TICs) which are found in both the sample and
       associated Wank(s^

6.     If an  instrument blank was  not  analyzed  following  a sample  analysis which-
 '      contained an anaiyte(s) at high concentrations), sample analysis results after the
       high concentration sample  must be evaluated for carryover.   Professional
       judgment should be used to determine if instrument cross-contamination has
       affected any positive compound identifications).   If instrument cross-
       contamination is suggested, then this should be noted  for TPO action if the
       cross-contamination is suspected of having an effect on the sample results.

7.     Blanks or samples run after a-matrix spike or  standard  should be carefully examined
       to determine the occurrence  of instrument or  syringe  carry-over.   Since the
       efficiency of sample  transfer  can vary dramatically  according  to apparatus and
       operator techniques, professional judgment should be used in each case  to determine
       whether sample or blank results are .attributable to  carry-over.   Some  common
       examples are as follows:         ,

       •      Zero to  one percent syringe carry-over occasionally in BNA runs.

       •      Higher percentages of carry-over following BNA runs that are saturated.

       Sample results which are  possible artifacts of carry-over should be flagged  as
       unreliable (R).

8.     When there is  convincing evidence  that contamination is restricted to a particular
       instrument, matrix, or concentration level, the 5X/1QX rule will only be applied to
       compare contaminated blanks to certain associated  samples (as  opposed to  all
       samples in the case).  Some examples are as  follows:

-------
                                                        Procedure No.:  M2-SVOA
                                                                      Revision:  1
                                                                 Date:  06/30/1995
                                                                     Page 27 of 39

       •     Column bleed (siloxanes) may be localized to a particular instrument.

       •     Common  laboratory  contaminants,  such  as  methylene  chloride  and
             phthalates, are generally too unpredictable to safely assume contamination is
             restricted to a particular instrument, matrix, or concentration level.

The following are examples  of: applying the blank qualification guidelines.   Certain
circumstances may warrant deviations from these guidelines.       "      .    :.     ,
      Example 1:   Sample result is  greater than the Contract Required Quantitation
                   Limit (CRQL), but is less than the 5x or lOx multiple of the blank
                   result. ••-'-•  '•' .. : .  '•',•- '           . ..  "  -••• •'•;'•;.  •-'•••-....'v:'-''. :-.'-•• '.-'.'
                                                            Rule
                          Blank Result
                          CRQL
                          Sample Result
                          Qualified Sample Result
                                                                5x
 '5,   5'
 60   30
 60S  30B
                   In the example for the "lOx" rule, sample results less than 70 (or 10
                   x 7) would be qualified "B".  In the case of me "5x" rule, sample
                   results less than 35 (or 5 x 7) would be qualified "B".
      Example 2:   Sample result is less than CRQL, and is also less than the 5x or lOx
                   multiple of the blank result.
                         Blank Result
                         CRQL
                         Sample Result
                         Qualified Sample Result
  Rule

IQx   5jc

 66
 5     5
 4J   4J
 4B  4B
                   Note that data are reported as 4B, indicating that  the qualitative
                   presence is not confirmed.

-------
                                                        Procedure No.:  M2-SVOA
                                                                      Revision: 1
                                                                 Date:  06/30/1995
                                                                     Page 28 of 39

       Example 3:   Sample result. is greater than the 5x or lOx multiple of the blank
                    result.

                                                                   Rule
                                Blank Result               10    10-
                                 Sample Result                  - 120   60
                                 QuaUfied.Sample Result           12(te  60
                    For both the "lOx" and "5x" rules, sample results exceeded the
                    adjusted blank  results  of -100 (or  10x10) and  SO  (or 5x10),
                    respectively. ^   .': "•-••' .: :-  -.'   : .    •-;-•.••'••..;'" :' '''•'•-.'•

3.7 Surrogate Spikes               •

Laboratory performance on individual samples is established by means of spiking activities.
All samples are spiked with surrogate compounds prior to sample preparation.   The
evaluation of the results of these surrogate spikes is not necessarily straightforward.  The
sample itself may produce  effects because of  such factors as interferences  and  high
concentrations of analytes. Since the effects of the sample matrix are frequently outside the
control of the laboratory  and may present relatively unique problems, the evaluation and
review of data based on specific sample results is  frequently subjective and demands
analytical  experience and professional judgment.   Accordingly,  this  section consists
primarily of guidelines, in some cases with several optional approaches suggested.

3.7.1  Acceptance Criteria

1.    Surrogate spikes, 4 acid compounds (3 required and 1 advisory) and 4 base/neutral
      compounds  (3 required and 1 advisory) are added to all samples and  blanks to
      measure their recovery in sample and blank matrices.

2.    Surrogate spike recoveries for semivolatile samples and blanks must be within the
      limits specified on Form n SV-1 and SV-2.

3.7.2  Review Items

Form II SV-1 and SV-2 and chromatograms.

-------
                                                         Procedure No.: M2-SVOA
                                                                        Revision:  1
                                                                  Date:  06/30/1995
                                                                      Page 29 of 39
3.7.3  Evaluation Procedures
 1.     Review chromatograms to check the surrogate spike recoveries on the Surrogate
       Recovery Form n SV-1 and SV-2.    •

2.     The following should be determined from the Surrogate Recovery form(s):

       a.      If any two base/neutral 01 acid surrogates are out of specification, or if any
             •one base/neutral or aad extractable surrogate has a recovery  of less than
              10%, then there should be a reanalysis to confirm that die non-compliance is
              because of sample matrix effects rather than laboratory deficiencies.

NOTE:       When meie are                                followed by successful re-
                     ^1he : laboratories are required to report only ;fhe successful run.
       D.    The laboratory has failed to perform satisfactorily if surrogate recoveries are
      .   . .;  out of specification and there is no evidence of reinjection of the extract, or
           :  reextraction and reanalysis (if reinjection fails to resolve the problem).

       c.    Verify mat no blanks have surrogates recoveries outside the criteria.
                  '.-•'"••.   •;      '     *      .        .           ' '
3.     Any time there are two or more analyses for a particular fraction the reviewer must
       determine which are the best data to report.  Considerations should include but are
       not limited to:

       a.    Surrogate recovery (marginal versus gross deviation).

       b.    Technical holding times.

       c.    Comparison of the values of the target compounds reported in each fraction.

       d.    Other QC information, such  as performance of internal standards.

4.    When both  the initial analysis and the reanalysis have surrogate recoveries outside
       of criteria,  the  data summary should normally  contain the highest concentration
      obtained for each compound  detected,  provided that surrogate recoveries  in  the
       analysis  being reported do not suggest a high bias.  However, if a demonstrated
      laboratory contaminant is detected in one analysis  but not the other, the negative
       result  may be more appropriate to report.

      When the  reanalysis  of a  fraction is  within  surrogate recovery  criteria,  the
      laboratory is required to provide only data for the acceptable analysis. If both sets
      of data are provided, and if a compound was detected in the initial analysis but not

-------
                                                         Procedure No-  M2-SVOA
                                                                       Revision:  1
                                                                  Date:  06/30/1995
                                                                      Page 30 of 39

       the reanalysis, then the positive result should be reported (provided the compound is
       not a demonstrated laboratory contaminant). The reported result should be flagged
       as estimated (J), due to possible sample inhomogeneity.

5.     If advisory surrogates are outside established criteria, professional judgment will be
       used in qualifying the sample results.  If the results are outside the criteria, then
       qualification would only affect similar target compounds..            '

3.7.4 Action

Data are not qualified with respect to surrogate recovery unless two or more semivolatile
surrogates, within the same fraction (base/neutral or acid fraction), are out-bf specification.
For surrogate spike recoveries out of specification, the following approaches are suggested
based on a review of all data from the case, especially considering the apparent complexity
of the sample matrix.    ;

NOTE:      These actions apply to all surrogates, except for  "advisory"  surrogates.
             Professional judgment should be used in qualifying sample results  based on.
             advisory surrogate recoveries.   Qualification based on advisory surrogate"
             recoveries  should  be applied to similar compounds  in the sample only.
             Specify in the .narrative  any actions taken based  on advisory surrogate
             recovery.                                                            .

1.     If two or more surrogates in either semivolatile  fraction (base/neutral or acid
       fraction) have a recovery greater than the upper acceptance limit (UL):

       a.     Specify the fraction that is being qualified; i.e. acid, base/neutral, or both.

       b.     Detected semivolatile target compounds are qualified biased high, "K".

       c.     Results .for  non-detected semivolatile  target compounds  should not  be
             qualified.

2.     If two or  more surrogates in  either semivolatile fraction have a recovery greater
       than or equal to 10% but less than the lower acceptance limit  (LL):

       a.     Specify the fraction that is being qualified, i.e. acid, base/neutral, or both.

       b.     Detected semivolatile target compounds are qualified biased low, "L".

       c.     For non-detected  semivolatile target compounds, the sample quantitation
             limit is qualified as biased low, "UL".

-------
                                                         Procedure No.: M2-SVOA
                                                                        Revision:  1
                                                                  Date:  06/30/1995
                                                                      Page 31 of 39

3.     If any surrogate in either semivolatile fraction show less than 10% recovery:

       a.    Specify the fraction that is being qualified, i.e. acid, base/neutral,  or both.

       b.    Detected semivolatile target compounds are qualified biased low, "L".

       c.    Non-detected semivolatile target compounds may be qualified as unusable
             (R).  (If advisory surrogate limits are not met, use professional judgment to
             qualify non-detected compounds).       :    :'-••'••:'•>•••'-.'••'•        ,   ,
... ::'.• --". ..-,•">... ;••-;. tabl
QUALIFICATION OF SEMI
SURROG;
. • • '.'..• '. ' •
Detected analytes
Non-detected analytes
2 or 3
All High
K
;None
CM2-SVOA
[VOLATJLE
ITE.BBCO1
2or3
All Low
L
UL
-5 ••'-•••••••:••••>'' ;•/.:•. . '•
V ANALYTES BASED ON
SERIES ::c,:-,-;>-\,.:.. ^%V '-'V '•
2 or 3
Mixed High/Low
••••-. :,J "--;:. ;••".;: '
ui
1 or More
< 10%Rec.
L
R
4.     If two or more surrogate recoveries in either semivolatile fraction (base/neutral or
      . acid fraction) are outside surrogate recovery limits, 'and one of the recoveries is
       below the lower limit (but > 10%) and the other recovery is above the upper limit:

       a.     Specify the fraction that is being qualified, i.e., acid, base/neutral, or both.

       b.     Detected semivolatile target compounds are qualified as estimated, "J".

       c.     Non-detected  semivolatile  target compounds are  qualified as  estimated,
             "UJ".

5.     In  the special  case of a blank  analysis with  surrogates out of specification, the
       reviewer must  give special consideration to the validity of associated sample data.
       The basic concern is whether the blank problems represent an isolated problem with
       the blank alone, .or whether there is a fundamental problem with the analytical
       process.  For  example,  if one or more samples  in the batch show  acceptable
       surrogate recoveries, the reviewer may choose to consider the blank problem to be
       an  isolated  occurrence.  However, even if this judgment allows some use of the
       affected data, analytical problems should be noted for TPO action.  Also note if
       there are potential contractual problems associated with the lack of re-analysis of
       samples that were out of specification.

-------
                                                         Procedure No.:  M2-SVOA
                                                                       Revision:  1
                                                                  Date: 06/30/1995
                                                                      Page 32 of 39

6.     Whenever possible,  the  potential  effects  of the data resulting from surrogate
       recoveries not meeting the advisory limits should be  noted in  the data  review
       narrative.

7.     Positive results for compounds already flagged for blank contamination will not
       need a separate flag for surrogate recoveries.  However, these situations should be
       addressed in the narrative or the support documentation.

8.     When  dilutions  are  performed  which prevent -detection  of  BNA  surrogate
    ..  compounds;-the narrative or support documentation should indicate.that extraction
       efficiency/methodaccuracy cannot be verified.
          •  ' :.;'-;;.:-:;-::r;';'\-  ;   "  "...   ;-  '  '•'.   --:  •  ''•    ' '        ...'•'.-  . •  "  •'• '/-.    /
9.,     Although semivolatile  surrogate recoveries  cannot usually  be  correlated with
       specific analytes, m the followmg cases s
       a particular suro?g^:A;r';\  •:.;;;; ..-.'"• '• ••    .. '";!:'.•••   •'; •' "•' '  •'.'.'  r   •

       a^     When a semivolatile surrogate is the deuterated analog of a TCL analyte (for
             exmple]r^phen6l;and pheiibi), a  low recovery for the surrogate can be
          :   used  to; flag positive results and quantitation limits as biased low for the ^
             undeuterated analog:  (This applies  even if no other surrogates are outside
             criteria .or if omer  surrogates are biased high mstead of low.)

       b.     When dij-ierphenyl is biased low, positive results and quantitation limits for
             me heavier polyaromatic hydrocarbons (those  which eliite  starting with
             fluorathene) can be considered as biased low. (This applies even if no other
             surrogates are outside criteria or if other surrogates are biased high instead
             of low.)

       c.     When 2,4,6-tribromophenol is biased, low,positive  results  and quantitation
             limits for trichlorophenols and pentachlorophenol  can be considered  as
             biased low.  (this applies even if no other surrogates are outside criteria or if
             other surrogates are biased high instead of low.)

3.8  Matrix Spikes/Matrix Spike Duplicates

Data for matrix spikes/matrix spike duplicates (MS/MSD) are generated to determine long-
term precision and accuracy of  the analytical  method on  various  matrices  and  to
demonstrate acceptable  compound recovery  by  the laboratory  at  the  time  of sample
analysis.  These data alone cannot be used  to evaluate the precision and accuracy  of
individual samples.  However, when exercising professional judgment, this data should be
used in conjunction  with other available QC information.

-------
                                                       Procedure No.: M2-SVOA
                                                                     Revision:  1
                                                                Date:  06/30/1995
                                                                   Page 33 of 39
3.8.1 Acceptance Criteria
 1.     Matrix spike and matrix spike duplicate samples are analyzed at frequency of one
       MS and MSD per 20 samples of similar matrix.
               j               •       • •
2.     Matrix spike and matrix spike duplicate recoveries should be within the advisory
       limits established on Form m SV-1 and SV-2.
     '  • •  '        ..•••'.     '':- '    • '     '-..-••.•   -. •••   I.-''-        •;   •
3.     The Relative Percent Differences (RPDs) between matrix spike and  matrix spike
       duplicate recoveries should be within the advisory limits listed on Form HI SV-1
       and SV-2.  '   / •..';'   '-..••.  •'     "   '   '    -."   "'- -..    •''.''

3.8.2  Data Requirements and Data Retrieval

Forai IE SV-1 and SV-2 and chromatograms.

3.8.3  Evaluation Procedures

1.     Verify that MS and MSP samples were analyzed at the required frequency and that
       results are provided for each  sample matrix.

2.     Inspect results for the MS/MSD Recovery on Form ffl SV-1 and SV-2 and verify
       that the results for recovery and RPD are within the advisory limits'.?

3.     Compare results (%RSD) of non-spiked compounds'between the original result,
       MS, and MSD,

3.8.4  Action

1.     No action is taken on MS/MSD data alone.  However, using informed professional
      judgment the data reviewer may use the matrix spike and matrix spike duplicate
       results in conjunction with other QC criteria and  determine the need for some
       qualification of the data.

2.     The data reviewer  should first try to determine to  what extent  the results of the
       MS/MSD affect the associated data.   This determination should be made  with
       regard to the MS/MSD sample itself as well as specific analytes for all samples
       associated with the MS/MSD.

3.     In those instances where it can be determined that the results of the MS/MSD affect
       only the sample  spiked, then qualification should be limited  to this sample alone.
       However, it may be determined through the MS/MSD results that a laboratory is

-------
                                                       Procedure No.:  M2-SVOA
                                                                    Revision:  1
                                                               Date: 06/30/1995
                                                                   Page 34 of 39

       having a systematic problem in the analysis of one or more analytes, which affects
       all associated samples.                          :       .

4;     The reviewer must use professional judgment to determine the need for qualification
       of positive results of non-spiked compounds.

NOTE:      If a field blank was used for the MS/MSD, a statement to that effect must be
             included for TPO action on the ORDAS.           ;,         -

5.     When extremely low  % recoveries are noted, qualify data for all affected
       compounds using professional judgment

6.     When non-spiked compounds are present in either.the MS or MSD results, a table
       in the data  review narrative is constructed.- showing original' (unspiked) sample
       results for non-spiked compounds, non-spiked compounds present in the MS and
       MSD and the calculated %RSD.

3.9  Internal Standards .-. •.";'*'
                        "     •" \" • • ,

Internal Standards (IS) performance criteria ensure that GC/MS sensitivity and response are
stable during every  analytical run.

3.9.1  Acceptance Criteria                                    ;^.

1.     Internal standard area counts for samples and blanks must not vary by more than a
       factor of two' (- 50% to + 100%)  from the associated calibration standard.

2.     The retention time of the internal standards in samples and blanks must not vary by
       more than +. 30 seconds from the retention time of the associated  calibration
       standard.

3.9.2  Review Items

Form Vin SV-1 and SV-2 and chromatograms.

3.9.3  Evaluation Procedures

1.     Check raw data (e.g., chromatograms and quantitation lists) for samples and blanks
       to verify the internal standard  retention times and areas reported on the Internal
       Standard Area Summary (Forms Vm SV-1, VHI SV-2).

2.     Verify that all retention times and  IS areas are within the required criteria.

-------
                                                         Procedure No.-:  M2-SVOA
                                                                       Revision:  1
                                                                  Date:  06/30/1995
                                                                     Page 35 of 39-

3.     If there are two analyses for a particular fraction, .the reviewer must determine
       which are the best data to report.  Considerations should include:

       a.     Magnitude and direction of the IS area shift

       b.     Magnitude and direction of the IS retention time shift.

       c.     Technical holding times.                         •

       d.     Comparison of the values of the target compounds reported in each fraction.

3.9.4 Action
                ,  ' '                    .           \                 '   •  .
                    '   '     • •  .  .                      •     ' .    .'••'•.'-.'    '*
1.     If an IS area count for a sample or blank4s*outside • 50% or + 100% of the area
       for the associated standard:       '                          .-,.,..

       a.     Positive results for compounds quantitated using that IS should be qualified'
         .''•  with'TV.  ...   '-'   ;   '.-•'•.  •   -;.  ••      -'   ' -.'.  .';." ••   -••-..'....    V;

       b.     Non-detected compounds quantitated using ah  IS area count  greater than
             +100%  should be qualified with "UJ".

       c.     Non-detected compounds quantitated using an IS area count less than 50%
             are reported as the associated sample quantitation limit and qualified with
             "UJ".

       d.     If extremely low area counts are  reported,  or if performance  exhibits a
             major abrupt drop-off, then a severe  loss of  sensitivity is indicated.   Non-
             detected target compounds  should then be qualified as unusable (R).

2.     If an IS retention time varies by more than 30 seconds:

       The chromatographic profile for that  sample  must be examined to determine if any
       false positives or negatives exist.  For shifts of a large magnitude, the reviewer may
       consider partial or total rejection (R)  of the data for  that sample fraction.  Positive
       results should not need to be qualified with  "R" if the mass spectral criteria are
       met.

3.     If the internal standards performance  criteria  are grossly exceeded, then this should
       be noted for TPO action.  Potential effects on the data resulting from unacceptable
       internal standard performance should  be noted in the data review narrative.

-------
                                                       Procedure No.: M2-SVOA
                                                                     Revision:  1
                                                               Date:  06/30/1995
                                                                   Page 36 of 39
3.10 Reported CRQLS
The objective is to ensure that the reported quantitation results and Contract Required
Quantitation Limits (CRQLs) for semivolatile target compounds are accurate.

3.10.1 Acceptance Criteria

1.     The CRQL  must be calculated according to the correct equation and account for
       moisture content and dilution factor as appropriate.      -

3.10.2 Review Items                 \

Form ISV-1 and SV-2 and chromatograms.

3.10.3 Evaluation Procedures

1.     Verify that the CRQLs have been adjusted  to  reflect  all  sample dilutions,
       concentrations,  splits,  clean-up activities, and  dry  weight factors that are  not;
       accounted for by the method;          .                   ;  •           '
           •        .'".      --..-''   r.' .'.-.
3.10.4 Action    \

1.     If there  are any discrepancies found,  the laboratory  may be contacted by  the
       designated representative to obtain additional  information that could resolve any
       differences.   If a discrepancy remains unresolved,  the reviewer  must  use
       professional judgment to decide which value is the best value.   Under these
       circumstances, the reviewer may determine qualification of data is  warranted.
       Decisions made on  data quality should be included in the data review narrative.  A
       description of the reasons for data qualification and the qualification that is applied
       to the data should be documented in the data review narrative.

2.     Numerous or significant failures to properly evaluate and adjust CRQLs should be
       noted for TPO action.

3.     The reviewer must assure that any results in  error by more than 10 percent  are
       identified and corrected  on the sample data summary.  If laboratory resubmission is
       not performed,  the reviewer should document his/her changes to the  data in  the
       narrative or  support documentation.  Calculation errors should also be noted on the
       ORDA.

-------
                                                        Procedure No.:  M2-SVOA
                                                                      Revision:  1
                                                                 Date:  06/30/1995
                                                                     Page 37 of 39
3.11  Tentatively Identified .Compounds
Chromatographic peaks in semivolatile fraction  analyses that are not target analytes,
surrogates, or internal standards are potential tentatively identified compounds  (TICs).
TICs must be qualitatively identified by a National Institute of Standards and Technology
(NIST) mass spectral library search and the identifications assessed by the data reviewer.

3.11.1

For each sample, the laboratory must conduct a mass spectral search of the NIST library
and report the possible identity for the 20 largest semivolatile fraction peaks which are not
surrogate, internal standard, or target compounds, but which have area or height greater
than 10 percent  of the area or height of the nearest internal standard.  TIC results are
reported for each sample-oS'liie^rganic Analyses Data Sheet (Form I SV-TIC).
             "~''     '              "      '   ''••
NOTE:      ;Smce              abn of October 1986, the CLP does not allow the
             laboratory tb'reporf is tentatively identified compounds any target compound
                1 •  • .-- . '^ —•;•.-••.'.ป-"*"••• '•"*4ปl'-ป.Vป.t, • p    '            •             *      ^^ •  ' '   ~
             which /is pro^^ly^r^rted in anomer fraction^  For example,  late eluting
             volatile target compounds should not be reported as semivolatile TICs.
              •  .'  '..)'•.- •'".: ''•:   '-.-•'.        '                 '      • '
                  . -   '   •  x ' ' '  *
'3.11.2  Review Items   <:-
               •',•'•"  "   '"•'"••

Form I SV-TIC, chromatograms.

3.11.3  Evaluation Procedures

1.     Guidelines for tentative identification are as follows:

       Ensure that TIC results are reported on Form Fs.

2.     Blank chromatograms  should be examined to verify that TIC peaks present in
       samples are not found in blanks. When a low-level non-target compound that is a
       common artifact or laboratory  contaminant is detected in a sample,  a thorough
       check of blank chromatograms may  require looking for peaks which are less than 10
       percent of the internal standard height, but present in the blank chromatogram at a
       similar relative  retention time.

3.     The reviewer should be aware of  common laboratory artifacts/contaminants and
       their sources (e.g.,  aldol condensation products, solvent preservatives, and reagent
       contaminants).  These may be present in blanks and not reported as sample TICs.

       Examples:

-------
                                                         Procedure No,: M2-SVOA
                                                                       Revision:  1
                                                                  Date:  06/30/1995
                                                                     Page38of39

       a.     Common laboratory contaminants:   CO2 (m/z  44),  siloxanes (m/z 73),
             diethyl ether, hexane, certain freons (l,l,2-trichloro-l,2,2-trifluoroethaneor
             fluoro-trichloromethane), and phthalates at levels less than 100 ug/L or 4000
             ug/Kg.

       b.     Solvent preservatives, such as cyclohexene which is a methylene chloride
             preservative.  Related by-products include cyclohexanone, cyclohexenone,
             cyclohexanol, cyclohexenol, chlorocyclohexene, and chlorocyclohexanol.

       c.     Aldol  reaction  products of  acetone  include:    4-hydroxy-4-methyl-2-
             pentanone, 4-methyl-2-penten-2-one, and 5,5-dimemyi-2(5H)-furanone.

4.     Occasionally, a target compound may be identified as a TIC in the proper .analytical
       fraction by non-target library search procedures, even though it was not found on
       the .quantitation list  If the total area quantitation method was used, the reviewer
       should request that the laboratory recalculate the result using the proper quantitation
       ion.  In addition, the reviewer  should evaluate  other sample chromatograms and
       check library reference retention times on quantitation lists to determine whether the
       raise negative result is an isolated occurrence or whether additional data may be •
       affected.

5.     Target compounds may be  identified in more than one  fraction.   Verify  that
       quantitation is made from the proper fraction.

3.11.4 Action

1.     All TIC results should be qualified "J", estimated concentration on the Laboratory
       Forml-TICs.

2.     Blank Results
                                                             /
       Form I-TIC which  contain sample results that  are questioned by blank results,
       should be flagged "B" and a line drawn through these data for emphasis (initialed
       and dated).
                   ',:
       To be considered questionable, a sample TIC concentration must be within 10 times
       the concentration of one of the blank results.  If different volumes/weights are used,
       the total amount of compound in the extract must be  compared for sample versus
       blank. In general, blanks analyzed within the same case, by the same lab, may be
       cross-applied to either soil or water samples extracted or analyzed on other days.

       All blank results must be attached in the support documentation section of the data
       review.

-------
                                                        Procedure No.:  M2-SVOA
                                                                      Revision:  1
                                                                 Date: 06/30/1995
                                                                     Page 39 of 39-

3.    When a compound is not found in any blanks, but is a suspected artifact of common
      laboratory contamination, the reviewer should cross off the reported TIC result on
      the copy of the Form I-TIC and note the reason(s) in the narrative.

4.    Physical constants,  such  as boiling point, may be  factored into professional
      judgment of TIC results.

5.    Failure to properly evaluate and report TICs should be noted for TPO action.

-------
                                                       Procedure No.: MI-PEST
                                                                    Revision:  1
                                                               Date:  06/30/1995
                                                                    Page 1 of 20:
                                 Appendix C-l
           Validation of Pesticides and  PCB Analyte Data
                            Manual Level Ml
                        1. Purpose and Applicability

This procedure provides step-by-step instructions to manually validate pesticide and ploly-
chlorinated biphenyl (pest/PCBs) data using the manual innovative data validation approach
atLevelMl.

The procedure  is applicable to the Pesticide/PCB  xlata  obtained using the  Conract
Laboratory Program Statement of Work (CLP SOW).  Hard copy data conforming to the
applicable CLP'SOW specifications are essentialin order to carry out .the procedure.  |

Data validated using  this procedure  are considered  usable  for the following  types of
purposes;  however, the  data users must decide on a case-by-case basis whether the
procedure is suitable for their intended data uses. The suggested data uses are:

      •      Oversight of activities led by other parties

      •      Action level comparison

      •      Initial site investigation

      •      Contamination sources


                  2.  Quality Control Measures Checked

Table MI-PEST-1 highlights the  quality control (QC)  indicators evaluated  under this  data
validation procedure.


                                3.  Procedure

The following subsections describe  for each  of the QC  indicators the acceptance criteria,
location and retrieval of QC data, evaluation  of the QC data, actions taken  in the event the
QC acceptance  criteria  are exceeded,  and  documentation of the QC  violations  in  a
standardized report form.

-------
Procedure No.:  Ml-PEST
            Revision:  1
       Date: 06/30/1995
            Page2o'f20
Table Ml-PEST-1
QC CHECKLIST FOR LEVEL Ml
CLP RAS PESTICIDES/POLYCHLORINATED BIPHENYLS
QC MEASURES
Action Level Notification
GC/ECD Performance Check '
Initial Calibration (CF) .
Initial Calibration (%RSD)
Continuing Calibration (RPD)
Laboratory Blank .
MS/MSD(%R,JRPD)
Field Quality Control (dup., blnk., PE samp.)
Sample Paperwork ; .
Holding Time
Retention Time
Surrogate Recovery
Dilution Factor • '
Moisture Content
Pesticide Cleanup Checks
Chromatograms
"
Raw Data
-'.. ./;../;•'. .- r Mamial;'/;:-
| Ml | ป0 W
r^!^
*?.'. .' ' •"•'
t •ป •*.

i
1 • . x |


i 1
1 ^
1 	 '
1 :• ' :..•'••-•
p :^v-
••.•.••v':-.-x--V-"'ir
- ••' •.•.ฃV) ••••;:*.
.. :.. '*.?.-v: .-.:
••'.-''•.x;r ••• '
''.'•X'": '•';-.'
/.•"x-'-V
1- •:••••• '.'
1 .'x .••
x ^
\". x."
1 - 'X .
X
x
X

X
X
I
•/•'". ' / = •
:;-.-;;M3 ; •
••••• x
'• •' • . i-
';':'." x r '
• ,.-:-.X "•
s-;-.-x '
•::-.'• -X •';
• ' X
!
X
X
X
X
X
X .
X .
X

x
X

-------
                                                         Procedure No.:  Ml-PEST
                                                                       Revision:  1
                                                                 Date:  06/30/1995
                                                                      Page 3 of 20
Reporting requirements for Level Ml are:

       •     Hand annotate the Form Fs, including
                    Data validation qualifiers
                    Sample identification number
                    Sampling location

       •     Provide a narrative that includes
             -      A statement that defines the level of the data review, i.e., Ml
                    Major problems associated with analysis

       •     Include the following attachments
                    List of data validation qualifiers
          '   -      Support documentation including forms that support assigning data
                    qualifiers
             -      Chain of custody form
         -    .    ~   " f       . m
3.1  Action Level Notification

The purpose behind action level notification is to make the EPA Remedial Project Officer
(RPM) or the  Site Project Officer (SPO)  aware of the potential human health risk at the
site.  In accordance with the Region HI Hazardous Waste Division policy, the EPA IJJPM
or SPO must be promptly notified of any contaminant exceeding the established action ijevel
or the 10-day health advisory limit. The data for contaminants exceeding the action levels
must be validated as a top priority and reported to the RPM or SPO.  Validation of the: rest
of the data may then be completed normally.

3.1.1  Acceptance Criteria

Region HI Hazardous Waste Division has established 10-day advisory limits or the action
levels for several organic  compounds and elements of special health risk concern.  The
pesticide organic compounds and their 10-day health advisory limits apply only to aqueous
samples and are listed in Table Ml-PEST-2.  The criteria for action level notification are
as follows:

       •     The contaminant concentration must be equal to or above the established 10-
             day health advisory limits.

       •     Data for contaminants exceeding the action levels must be validated as a top
             priority.

       •     The  following EPA  personnel  must  be  notified  of  the action level
             exceedances:

-------
                                          Procedure No.: MI-PEST
                                                       Revision: 1
                                                  Date: 06/30/1995
                                                       Page 4 of 20
      EPA RPM or SPO

      EPA Section Chiefs:

      -     Site Investigation (SI)

      —     Remedial

      -     Enforcement

             RCRA
      (
      EPA Section Toxicologists:^ .

      -     Enforcement

      —     Superfund

     —-.    RCRA

The remaining data validation should be completed per normal procedures.

Any  special  instructions  from the Hazardous Waste Division should be
followed.

Records should be kept of the data review, action level notification and any
follow-up instructions and actions.
Table Ml-PEST-2
PESTICIDES AND POLYCHLORINATED BIPHENYLS AND ACTION
LEVELS
Compound
Chlordane
Heptachlor
Methoxychlor
Action
Level*
63
10
2,000
Compound
Endrin
Lindane
Toxaphene
Action
Level*
5
1,200
80
*A11 units are ug/1.

-------
                                                         Procedure No.:  Ml -PEST
                                                                      Revision:  1
                                                                 Date:  06/30/1995
                                                                     Page 5 of 20
 3.1.2  Review Items
All data required to perform the complete Level Ml validation, as detailed in the following
sections are necessary for carrying out action level notification.  The location of the data
.and their retrieval procedures are also discussed below.

3.1.3  Evaluation Procedure

The evaluation process  preceding  action level  notification will  primarily consist  of
comparing the results on Form I*s with the action levels presented in Table Ml-PEST-2.
Following the identification of the contaminants exceeding the action levels, focused data
validation 'should  be performed  using the criteria, and procedures  described  in 1 the
appropriate sections below.      :             •*•"••

3.1.4  Action

The action resulting  from focused data validation will be the notification of action level
exceedance to the personnel identified above in Section 3.1.1.   :                   \
                           \            .               •           '         ,
3.1.5  Reporting

Copies of Form I's can be used to highlight the contaminants above the action levels. The
findings of the focused validation can be summarized  in  a  memorandum, and the data
qualifiers  resulting from focused validation may be written on the Form I's.  The marked
up forms should be clarified  that  they represent validation of only the contaminants
exceeding the action levels,  and not all data.

3.2 Evaluation of Retention Times

Retention  times are  the  only tool (in the event  that a mass spectral analysis was not
required)  that allow for the identification of the pesticides and PCBs.  While a retention
time from a single column is not  an unequivocal proof of a compound's presence, if the
retention  time of the suspected compound on a second column also matches  that. of a
standard,  then the compound's presence is  deemed  confirmed.   There  is reasonable
probability that a  non-target compound  may have the same retention  time as a  target
compound on one gas chromatographic column, but the probability of the two compounds
having  same  retention times also  on a second column is indeed very remote.  For this
reason,  the methods utilizing non-specific detectors require that the analyses be performed
under two separate sets of chromatographic conditions

-------
                                                         Procedure No.: MI-PEST
                                                                      Revision:  1
                                                                 Date:  06/30/1995
                                                                     Page 6 of 20
3t.2.1  Acceptance Criteria

Tjhe retention time acceptance criteria have been established by the EPA and Region m as
follows:

      The retention times of both of the surrogates, matrix spikes, and reported
      compounds in each  sample must be within the calculated .retention  time
      windows on both columns.  The acceptable windows are ฑ0.05 minutes for
      compounds eluting  before  heptachlor epoxide  and  ฑ0.07  minutes for
      compounds eluting after  and including heptachlor epoxide.. The retention;
      time for the surrogate tetrachloro-m-xylene (TCX) must be within ฑ0.05
      minutes of the mean retention time determined from  the initial calibration
      and that for decachlorobiphenyl (DCB) must be within  ฑ0.10 minutes of the
      mean retention time determined from the initial calibration.

3.2.2  Review  Items

The instrument level  printouts  or the quantitation  reports  are required  to obtain  the
retention times for the detected compounds. These are included in the raw data sections of
the data package  for the calibrations as well as the samples. Additionally, Forms Vm and
X would be helpful in substantiating and documenting any discrepancies.  It also may be
helpful to use a copy of the necessary calibration quantitation reports for a .comparison with
the sample retention times.

3|.2.3  Evaluation Procedures

Note: It is important to emphasize that the evaluation of the retention times goes hand-in-
hand with the chromatographic evaluations. During the evaluation  of any one of these QC
measures,  a  substantial use of the other QC measure is involved.  Therefore,  it may be
beneficial to carry out these two evaluations together.

      •     Review the quantitation reports and chromatograms for retention times in the
             standards  and samples.  Use Forms I, VIII and X as additional tools  for
             documenting  and  confirming the accuracy of  the reported data.   Confirm
             reported detected  analytes  by comparing the sample chromatograms to  the
             tabulated  results  and verifying peak  measurements and retention times.
             Acceptable ranges for the pesticides and PCBs are presented in Table Ml-
             PEST-3.  Ensure that the sample retention times are within these ranges for
             the compounds reported as detected and  confirmed by  the second column
             analysis.  Perform a similar evaluation of the associated blank data for  a
             confirmation of the reported laboratory contaminants, if any.

-------
                                                         Procedure Nq.:  MI-PEST
                                                                      Revision:  1
                                                                Date:  06/30/1995
                                                                     Page 7 of 20:
       •      For  multi-component  target compounds (Toxaphene and  Aroclors), the
             retention times and relative peak height ratios of major component peaks
             should be compared against the'appropriate standard chromatograms.

3.2.4 Action

If the qualitative criteria for both columns were not met, all: target compounds that' are
reported detected should be considered riondetected.  The reviewer may need to use the
qualifiers that are specific to pesticides. The reviewer should use professional judgment to
assign an appropriate quantitation; limit using the following guidance..               ;
                   •  •   .      .-  • •   •    •'  -   .''•.'      ':.-''"•'••'.    .     j
       •      If the misidentified  peak  was sufficiently  outside the  target pesticide
             retention time'window, then the repotted values may, .be a false positive; and
             should be replaced with the sample CRQL value.  V  .    .      ;    :

       •      If the misidentified peak poses an interference with potential detection of a
             target peak, the reported  value should be  considered and  qualified  as
             unusable (R).    .                     -.  '",            . :
                                              " .            •            .     \
       •      If the data reviewer identifies a peak in both GC column .analyses that falls
             within the  appropriate retention time  window,  but was  reported  as  a
             nondetect, then  the compound may  be a  false negatives-  Professional
             judgment should be used to decide if the compound should be included.

             All  conclusions made regarding target compound; identification should be
             included in  the data review narrative.
                                                                             i
3.2.5 Reporting

Include one or more Forms Ml-PEST-WNDW (example provided by Table Ml-PEST-4)
to represent the acceptance windows for  the retention times based on the  calibrations.
Record any deviations of the retention times on the  Form Ml-PEST-RT for the detected
compounds in each sample.  Include these  documents in the report for Level Ml data
validation results.   If  the  data need  qualification,  enter appropriate  qualifier code on the
sample Form  I's, and attach  these to the report.

3.3  Evaluation of Chromatograms For Detected Compounds

The purpose behind evaluating the chromatograms is to get an idea regarding potential false
negatives, and gross analytical  errors. Some idea as to the false positives  may also be
derived by checking the chromatograms.  Evaluation of the positive data as.described under

-------
                             Case No.:_
                             SDG No.:.
                   Data Reviewer:	
                            Site:
 Procedure No.: Ml-PEST
        •      Revision:  1
	  Date:  06/30/1995
             Page 8 of 20
Oatei Analyzed:
                                    Instalment ID:
Table Ml-PEST-3
. Pesticides/Polycblorinated BipUenyls Retention time Windows
Compound
•Iplu-BHC . ,. " '•.:.-.
beta-BHC . -
deb-BBC
gammt-BHC (Liodane)
Heptichlor . .
Heptachlorepoxtde '•- -'•<•
ftiiitaftflfin I • •:'.-. • * • -
lyifffflpH • _ • " '• • ^ ",-'.'•
4,4'-DDB . ••':-. . - .;
Emfriii ' - ' . ''•''- '
VnAft**ttfm9* If ' ' " ' - • ' . '

4,4*-DDD : '
findbtul&n fulftte . . -
4,4'-DOT
Methoxyehlor
Eodrio ketone. .•
Eodrin aldehyde
alpha-CUordane
gamma-Chlordane
Toxaphene
Aroclor 1016
Aroclor 1221
Aroclor 1232
Aroclor 1242
Aroclor 1248
Aroclor 1254
Aroclor 1260
Tetrachloro-m-xylene (Surr.)
Decachlorobiphenyl (Surr.)
Retention Time
Window in
, Mianttt
ฑ0.05
ฑ(WJ5
:ฑOM
ฑ0.05 .
. ฑ0.05
ฑ0.07
. ฑ0.07
:ฑd.0ฃ:
^•••"tWito"'.-
. ฑo:o7
" J.Aft7

ฑ0.07
ฑ0.07-
ฑ0.07
ฑ0.07.
ฑ0.07
ฑ0.07
ฑ0.07
ฑ0.07
ฑ0.07
ฑ0.07
ฑ0.07
ฑ0.07
ฑ0.07
ฑ0.07
ฑ0.07
ฑ0.07
ฑ0.05
ฑ0.10
Primary Column
Sid. Ret. Time





•

/ . • -




':








-

-

'



Range
* . ;
"




,






















Secondary f"i^"""
SUJ. Ret. Time

•• . •


' • .-.''•:.'










. - ^_
. . • •...












Range
• •, ' 1 '
i

i •


\.
• \




1
!
' '
I















-------
         Case No.:_
         SDG No.:.
Data Reviewer:	
        Site:	
 Procedure No.:  MI-PEST
             Revision: 1
	   Date:  06/30/1995
            Page 9 of 20
             -Table Ml-PEST-4
PEST/PCB Retention Time Evaluation Summary
Dates Analyzed:
/•'
i
• Inftn'uiftn1 IDs
. ' Column (1): '
. . • Column (2): •
Compound: • . -
alpha-BHC
beU-BHC
delta-BHC
gunma-BHC (lindane) .
Hepuehlor
Heptaehlor epoxide
EndosulfanI
Dieldrin
4,4'-DDE
Endrin
EndosulfanJI
4.4MJDD
Endosulfan sulfate
4,4'-DDT
Methoxychlor
Endrin ketone
Endrin aldehyde
alpha-chlordane
gamma -chlordane
Toxaphene Peak 1
Peak!
Peak 3
Aroclor 1016 Peak 1
Peak 2
Peak 3
Aroclor 1221 Peak 1
Peak 2
Peak3
Aroclor 1232 Peak 1
Peak 2
Peak 3
Aroclor 1242 Peak 1
Peak 2
Peak 3
Aroclor 1248 • Peak 1
Peak 2
Peak 3
Aroclor 1254 Peak 1
Peak 2
. Peak 3
Aroclor 1260 Peak I
Peak 2
Peak3
Sample Identifier:
1.
2.
3.
4.. ' - :
5. • .• • '
6. •..'.;.'
7, • • ; : '••
8. " . '•''••..
9. • - • ••' '
10.

Comments :
ป• •
2.
3. . . . • .,..•-
4. . ' ' ••
5. • • ' . ' ,
ซ• • • . • • -• ,- . . i-
7; ' ' ' ' ' ._.''>•
-*. . ' . • . ' ' .•••'• •..••. •
9. ' . ...-.-•...-.
10.

Sample Identifier & Retention Times for Primary Column (1) and Secondary Column (2)
1
a)








.


































a)

..









































2
0)











































C)
,• ..










































3
tt)
>










































0)











































4
a)











































w











































5 .
av






























"






























































10 .
tt)











































0)



.
%







































-------
                                                          Procedure No.:  Ml -PEST
                                                                       Revision:  1
                                                                  Date: 06/30/1995
                                                                      Page 10 of 20
mass spectral evaluation does not offer any insight into data that are not reported (i.e.,
reported  as nondetects).  Laboratory error or gross interference from other compounds
cbuld be the reasons for erroneous reporting.  The gas chromatographic profiles are the
primary tool used for the false negative evaluation under Level Ml.

$.3.1  Acceptance Criteria

TJhere are no EPA.-established quantitative criteria for evaluating chromatograms for false
negatives.  The criteria used for evaluating chromatograms are based on good laboratory
and scientific practices, and these are not hard and fast requirements.  The suggested.
evaluation criteria are as follows:
                    ':''.*•..•'''•-.''                   "     .         '
       •     Chrpmatpgrams must displaysingle  component pesticides  detected in the
             sample and the largest peak of any multi-component analyte detected in the
             sample at less than full scale,                          •          '.'•-.'

       •     If an extract must be diluted, chromatograms must display single component
             pesticides between 10 and  100 percent  of full scale, and multicomponent
             analytes between 25 arid 1(X) percent of full scale.
             •       ''       "ซ.''--•     . -                   _       f •
       •     For any sample, the baseline of the chromatogram must return to below 50
             percent of full scale before the elution time of alpha-BHC, and;also return to
             below 25 percent of full scale  after the  elution time of alpha-BHC  and
             before the elution time of decachlorobiphenyl,

       •     If a chromatogram is replotted electronically to meet these requirements, the
             scaling factor used must be displayed on the chromatogram,and both the
             initial chromatogram and the replotted chromatogram must be submitted in
             the data package.

       •     There  should not be  any significant  peaks in the chromatograms that are
             accounted for as TCLs.  Significant peaks are those with a minimum peak
             height of 10 percent of the  full scale deflection.

       •     The chromatograms  should  ideally have  base-line  .resolution  between
             adjacent peaks. Also, there should not be broad (unresolved) envelops in the
             chromatograms.

       •     There should not be abrupt shifts in the baseline.

       •     There should .not be peak tailing or sharp rise in the peak fronts.

-------
                                                         Procedure No.:  MI-PEST
                                                                      Revision:  1
                                                                 Date: 06/30/1995
                                                                     Page 11 of 20
3.3.2 Review Items
Chromatograms for each sample and standard calibrations are necessary for the evaluation.
These can be found in the front of the raw data package for each sample.  The quantisation
report  for each  sample is also  necessary to  retrieve the retention  times  since I the
chromatographic profile from a single column can be inconclusive.  A copy of the standard
chromatograms from  both chromatographic columns may be  very helpful  in visually
comparing the sample: traces for fingerprint features.     .   .     •                 !

3.3.3  Evaluation Procedures

Note: It is important to emphasize that the evaluation of the chromatograms goes hancl-in-
hand with the retention time evaluations.  Duritlg the evaluation of any one of these
measures,  a substantial use of the other QC measure is involved.  Therefore, it ma
beneficial to carry out these two evaluations together.

             Visually inspect the primary and the secondary column chromatograms
QC
 be
for
             all peaks that appear to be at least .10 percent of the full scale deflection.
             Compare the retention time of the suspect compound on the primary column
             with that of the standard.  If the retention time is close to the acceptance
             range, then check the retention time on the secondary column.  If there is
             clear disagreement in the retention time the compound is likely not a target
             compound and should not have been reported.

             If multicomponent. target, compounds,  exhibit  marginal pattern-matching  '
             quality, professional judgment should  be used to establish whether  the
             differences are due to environmental "weathering" (i.e., degradation of the
             earlier eluting peaks relative to the later  eluting peaks).  If the presence of a
             multicomponent pesticide is strongly suggested, results should be reported as
             presumptively present (N).

             If an observed pattern closely matches more than one Aroclor, professional
             judgment should be used to decide  whether  the neighboring Aroclor is a
             better match, or if multiple Aroclors  are present.

             If GC/MS confirmation was required but not performed, the reviewer should
             report this for TPO action.

             Also  observe the  chromatogram traces for  peak resolution between  the
             adjacent single component peaks.  Poor peak-to-peak resolution is indicative
             of degrading performance of the gas chromatographic column.  The values
             obtained from a degrading system are prone  to be inaccurate.  Generally,
                                                            U.S. EPA Headquarters Library
                                                                  Mail code 3201
                                                            1200 Pennsylvania Avenue NW
                                                               Washington DC 20460

-------
                                                         Procedure NOJ  MI-PEST
                                                                      Revision:  1
                                                                 Date: 06/30/1995
                                                                     Page 12 of 20
             there should be at least 90 percent valley between the neighboring peaks.
             This criteria is important for detectors such as the electron capture detector
             (BCD) that do not allow for unequivocal identification. Poor peak resolution
             between the adjacent peaks may results .in estimated quantisation of both for
             both compounds.

             Inspect the BCD  .profile for broad,  unresolved envelops.   These  are
             generally indicative of outside interference  from homologous compounds.
             Such envelops  .can  mask  the target  peaks or interfere, with accurate
             quantitation of the peaks.                               >;.
       •      Inspect the BCD profile for abrupt shifts in the baseline.  Such shifts are
             indicative of problems with instrument sensitivity or leakage in the system.
             The area counts obtained from shifted baseline are inaccurate, or even the
             detection of a TCL at low concentration may be missed.
                    "•'.-'••.' i     .-\ '   "'     '          •'•''          .  .
       •      Rapid  peak  rising  or  peak tailing  indicate problems  with,  the  gas
             chromatographic column, such as depleted stationary phase  on the column,
             decomposition of the stationary phase or creation of active sites.  Again, a
             visual inspection, of the BCD profile will yield information on the shape of
             the peak.            ;                .                 '

3.3.4 Action                                                      ^

professional discretion  must be used when evaluating and qualifying  data based on the
chromatographic evaluations.  An experienced chemist can generally infer the magnitude
and the frequency of the problem from the ECD profile and fingerprints.  If the problem
appears to be systematic, then it  should be brought to  the laboratory's  attention  and
resolved.   Intermittent problems may  or  may not require  any  action.  The following
guidelines are suggested when acting on ECD. profile observations:

       •      Any unaccounted  TCL peak (as confirmed by  the retention times on both
             columns) with area equivalent to or greater than the lowest  reportable limit
             for the sample must be brought to the laboratory's attention and resolved.
             In the event, the  discrepancy cannot be resolved with  the laboratory, the
             problems should be documented  and brought to the attention of the CLP-
             TPO, the RPM and  the SM.   The  data for unreported  TCLs may be
             considered unusable until the problems are resolved.

       •      If a peak resolution problem is evident for the samples, and appears to be
             systematic (i.e., present in  all calibration samples,  QC  samples, and field
             samples and increasing as the run progresses, additional QC measures such

-------
                                             Procedure No.:  MI-PEST
                                                          Revision:  1
                                                     Date: 06/30/1995
                                                         Page 13 of 20
as the continuing calibration percent difference (%D), and surrogate standard
recoveries  in  the  vicinity  of the  affected peaks should  be evaluated to
determine  if  the  peak resolution  problem  could  affect  detection!  or
quantitation.   If determined so, the positive  data  may  be qualified as
estimated, "J." Negative data may also be qualified as estimated, "UJ
if
the ability to detect at low concentrations is also deemed to be jeopardized
by poor resolution of adjacent peaks.                      .              .

Broad envelopes of homologous compounds could interfere with quantitation
or even detection. If the interference is evident from the recoveries of the
surrogate standards in the vicinity of the envelop, associated compounds may
also be interfered with.   Using professional discretion, the  positive iand
negative data may be considered as estimated, "J-? and  "UJ", respectively.
If the project objectives cannot be met with the estimated  data, alternative
sample preparation and cleanup procedures may need to be developed land
specified. The recommended solution should be brought to the attention of
the RPM, the SM and the TPO.                                     j  • •

Discrete shifts in the baseline in the  middle of a run are  indicative  of
intermittent problems. If the shift is due to leakage or change in the system
pressure, the positive as well as negative data may be considered estimated
("J"  and "UJ,  respectively).   The problem could be also due to  sbme
fluctuation in the instrument electronics which may lead to drastic changes in
the sensitivity of the instrument to detect the compounds.   As  a  note,
professional judgment should be exercised in determining the severity of the
problem.  For example, the magnitude of  a drop in  the baseline below! the
zero line may not be estimated and could be very significant.  On the other
hand, a drop that yields  a baseline  still above zero can  be put in a
perspective with the original baseline and a general appearance of the entire
RIC profile.   .   '   ' .

The problems with peak symmetry are indicative,of system degradation, and
should be brought to the attention of the laboratory for  a corrective action.
.Professional judgment should be used when and if qualifying any data due to
unsymmetrical peaks.  First the  problem  should be defined in terms  of
persistence throughout the chromatogram and  also from sample to  sample.
Additionally, the shapes and area counts for the surrogate standards should
be evaluated to see if the problem could have affected compound detection
and/or quantitation.  Data qualification  may: be uncalled for if the standard
area counts are acceptable.

-------
                                                         Procedure No,:  MI-PEST
                                                                       Revision:  1
                                                                 Date: 06/30/1995
                                                                     Page 14 of 20
3.3.5 Reporting

Keeps notes of the problems and substantiate them with copies of both cnromatograms and
ojther pertinent laboratory paperwork. . Include these items in the overall report for Level
Ml data validation.  For more descriptive comments, a separate sheet may be used.

3.4  Evaluation of Blanks

"the  purpose  of laboratory (or  field) blank analyses is to determine the existence and
magnitude of contamination problems resulting from laboratory (or field) activities.  The
criteria for evaluation of laboratory blanks apply to any blank associated with the samples
(e.g., method blanks, instrument blanks, and sulfur cleanup blanks).  If problems exist with
a ly type of blank, all associated data must be carefully evaluated to determine whether or
not there is an inherent variability in the data, or if the problem is an isolated occurrence
not affecting other data.                                          '.,'•••     •   ',

laboratory blanks and field blanks have a profound impact on false positives  reported in
samples; i.e.,.compounds reported as positive detects but not originating from the samples
themselves.  Cross contamination from  the sampling equipment, incidental contamination
from  the field  conditions or contamination  from the laboratory equipment  or general
environmental are likely sources of false positives in the samples.

3.4.1 Acceptance Criteria   "•''..

Criteria for blank evaluation are specified in the EPA's functional guidelines.  In addition,
Region HI  has  some additional requirements modifying the guidance.  The  acceptance
criteria for blanks apply equally to any type of blanks associated with either sampling or
analysis,  such as trip blanks, rinsate blanks, field or bottle blanks,  laboratory method
blanks.   While  there are several  criteria for evaluating  the blanks, the only criteria
applicable to Level Ml is the comparison of the blank and .sample concentrations. These
criteria are as follows:

             No contaminants should be present in the blanks.

             Frequency of Blank analyses:

                   Method  Blanks—A method blank analysis must be  performed for
                   each  20 samples  of similar matrix in each sample delivery  group
                   (SDG) or whenever a sample extraction procedure is performed.

                   Instrument Blanks—An acceptable instrument blank must be run at
                   ; least once  every  12 hours and immediately prior to the analysis of

-------
                                                          Procedure No.:  MI-PEST
                                                                        Revision:  1
                                                                  Date: 06/30/1995
                                                                      Page 15 of 20
                                                                               i

                    either the performance evaluation mixture or Individual Standard
                    Mixtures A and B, depending on the place in the analysis sequence.

                    Sulfur Cleanup Blanks—A  sulfur cleanup' blank must be analyzed
                    whenever part of a set of samples extracted together requires sulfur
                    cleanup.  If the entire set of samples associated with a method blank
                    requires  sulfur cleanup,  then the method blank also serves: the
                    purpose of a sulfur blank and no  separate sulfur blank is required.

3.4.2 Review Items

Data requirements and data retrieval procedures for blanks are the same as those  for the
field samples because the blanks as well as field samples are validated similarly.  Form! I's,
mass  spectra, chromatograms, quantitation reports, etc., are essential  for  performing a
validation of the blanks first.                             '         .

3.4.3 Evaluation Procedure

Validate the blanks same as the field samples. Detailed validation procedures are described
above under appropriate sections.  Use the validated blank results for a comparison with
the sample results.  Make certain that the samples and blanks are evaluated on the same
basis  of sample weight or volume, dilution factors,  moisture content, eUC"~Use the 5 (or
10) times the highest blank concentrations for  qualifying the  sample  data.    More
specifically, the blank data evaluation procedures are as follows:

       •       Review the results of all associated blanks, Form I PEST, and Form IV
              PEST, and chromatograms to evaluate the presence of TCL pesticides.

       •       Verify that method blank analysis has been reported  per SDG, per  matrix,
              per concentration level, for each GC system used to analyze samples, and
              for each extraction batch.

       •       Verify that the  method blank analyses do not contain  any target pesticide or
              Aroclor/Toxaphene at greater than its Contract Required Quantitation Limits
              (CRQL).          -

       •   .    For the surrogates in  each method  blar\k, verify that  the observed retention
              times'are within the appropriate retention time windows calculated from the
              initial calibration.

       •       Verify that the instrument blank analysis has been performed every 12 hours
              as part of the continuing calibration and following a  sample analysis which

-------
                                                          Procedure No.: MI-PEST
                                                                       Revision:  1
                                                                  Date:  06/30/1995
                                                                      Page 16 of 20
             contains an analyte(s) at high concentration(s), and that the instrument blanks
             do not contain any target analytes above one-half the CRQL, assuming that
             the material in the instrument-resulted from the extraction of a 1-L water
             sample;
                                         \         /         .    "    /* .
             Verify that the sulfur cleanup blanks were analyzed at the required frequency
             and  that  they  do not contain any  target compound  above the  CRQL,
             assuming that the material in the instrument resulted from the extraction of a
             1-L  water sample.  If a separate sulfur cleanup blank was prepared, one
             version of Form TV PEST should be completed associating^ the  samples
             with the method blank, and a second version of Form IV PEST should be
             completed  listing only those samples  associated with  the separate sulfur
             cleanup blank.

3,.4.4 Action

If the appropriate blanks were not analyzed with the frequency described in Criteria 3, 4,
and  5, then  the data reviewer should use professional judgment to determine if the
associated sample data should be qualified. The reviewer may need to obtain additional
information from the laboratory.  The situation should be brought to'the attention of the
TPO.                     .                                   ,

Action in the case  of unsuitable blank results depends on the circumstances and the origin
of the blank.  Detected compound results should be reported unless  the concentration of the
compound in the sample is less  than or equal to 5 times (5x) the amount in the blank. In
instances where more than one blank is associated with a given sample, qualification should
be based  on a comparison with  the associated blank having the highest concentration of a
contaminant.  The  results must not be corrected by subtracting the blank value.

       Specific actions are as follows:

       •      If a  target pesticide or Aroclor/Toxaphene is  found in  the blank  but not
             found in the sample(s), no qualification is required.  If the contaminant(s) is
             found at level(s)  significantly greater than the CRQL, then this should be
             noted in the data review narrative.

       •      Any pesticide or  Aroclor/Toxaphene detected in the sample, that was also
             detected in any associated blank, is qualified if the sample concentration is
             less  than five times (5x) the blank concentration.  The quantisation limit may
             also  be elevated.    Typically,  the  sample  CRQL is elevated   to  the
             concentration found in the sample.  The reviewer should  use professional
             judgment  to determine if further elevation of the CRQL is required.

-------
                                                           Procedure No.: Ml -PEST
                                                                        Revision:  1
                                                                   Date:  06/30/1995
                                                                       Page 17 of 20
             The reviewer should note that analyte concentrations calculated for method
             blank may not involve the same weights, volumes, or dilution factors as the
             associated  samples.  These factors must be taken into consideration  when
             applying the "5x" criteria, such that a comparison of the total amount of
             contamination is actually made.
       •     In addition,  there may be instances  when little or no contamination |was
             present in the associated blanks, but qualification of the sample was deemed
             necessary.    Contamination introduced through dilution  is one example.
             Although it is not always possible to  determine, instances of this occurring
             can be detected when contaminants are found in the diluted sample re ;ult,
             but absent  in  the undiluted sample  result.    Since  both results  are not
             routinely reported, it may be possible'  to verify this source of contamination.
             However, if the reviewer determines that the contamination is from a so irce
             other than the sample, he/she should qualify the data.   In this case, the
             "5x"  rule  does not apply; the  sample value should be reported  as  a.
             nondetected target compound, "U."                                      ซ

       •     If gross contamination exists (i.e., saturated peaks), all affected compounds
             in the  associated samples  should be qualified as  unusable  (R), due to
             interference.   This should  be noted in the data review  narrative if the
             contamination is suspected of having an effect on the sample results.

       •     If inordinate amounts of other  target  pesticides or Aroclors/Toxaphene are
             found at low levels in the blank(s),, it  may be indicative of a problem at the
             laboratory and should be noted  in the  data review narrative.

       •     If an instrument blank was not analyzed following a sample analysis which
             contained an analyte(s) at high concentration(s), sample analysis results  after
             the high concentration sample must be evaluated for carryover.  Professional
             judgment should be used to determine if instrument cross-contamination1 has
             effected any positive  compound identification(s), and  if  so,  detected
             compound results  should be qualified.  If instrument cross-contamination is
             suggested, then this should be noted in the data review narrative if the cross-
             contamination is suspected of having an effect on the sample results.

The following  are  examples of applying the blank  qualification guidelines.   Certain
circumstances may warrant deviations from these guidelines:

-------
                                                          Procedure No.;  MI-PEST
                                                                       Revision:  1
                                                                  Date: 06/30/1995
                                                                      Page 18 of 20


       Example 1:   Sample result is greater than  the  CRQL, but is less than  the 5x
                    multiple of the blank result.
                          Blank Result                     1.0
                          CRQL                           0.5
                          Sample Result                    4.0
                          Qualified Sample Result           4.0B

                    In this case, sample results less  than 5.0  (or 5  x 1.0) would be
                    qualified as nondetected target compounds.
       Example 2:   Sample result. is less than the CRQL, and is also less than the 5x
                    multiple of the blank result.                              '
                          Blank Result                     LO
                          CRQL                           0.5
                          Sample Result                    0.4J
                          Qualified Sample Result           0.4B

       Example 3:   Sample result is greater than the 5x multiple of the blank result.

                                                           5jc

                          Blank Result                     1.0
                          CRQL          .0.5
                          Sample Result                    10.0
                          Qualified Sample Result           10.0

                    In this case, the sample result exceeded the adjusted blank result (5x
                    1.0) and the sample result is not qualified.

       •     Record blank contaminants on QSFs.

If the sample  concentration  do  not  meet the  criteria of  5  (or   10) times  the  blank
concentration, the sample results should be considered essentially  undetected (or as not
detected substantially above the levels reported in the blanks); therefore, flagged "B" in
accordance with the Region III data validation guidelines.

-------
                                                         Procedure No.:  MI-PEST
                                                                       Revision:  1
                                                                 Date:  06/30/1995
                                                                     Page 19 of 20
3.4.5 Reporting
Form I's may be used to write the "B" data'qualifier for the data not meeting the blank
criteria.  Additional discrepancies may be included in the overall report for the Level Ml
validation.

3.5  Sample Paperwork

The purpose for evaluating the sample paperwork is to determine that the samples being
validated are indeed the ones taken from the site, and have  not been tampered with.
Accurate sample  identity is of. paramount importance in substantiating  the sample data.
Without unequivocal sample identity and chain-of-custody procedures, the sample data may
hot be defensible or enforceable.               *• - . .               .'•.":
                     ''••.•••       '      ' •   •  .    ' ••'    '      -.:•'•-.••.•'!
Under the current CLP contracts, the original paperwork (i.e.,  the purge package or the
administrative record) is included in the data package from the laboratory.  It is assumed
that the data validator is not privy to the original paperwork; therefore, the evaluation"
criteria and procedures described  below apply only to the documents that are generally
included in the data validation package.  These documents are the chain-of-custody fqrms
and Region m Shipping Record.

3.5.1 Acceptance Criteria                           .           C

Criteria for acceptability or authenticity of the sampling paperwork, document control! and
chain-of-custpdy have been established by the National Enforcement Investigations Center
(NEIC), in support of the CLP.   Overall criteria are too numerous and subjective to be
discussed here, but the criteria that apply to data validation are:

       •      The chain-of-custody form should be  properly and completely  filled out
             including the sampler signatures, date and time of sampling, sampling station
             identification, analyses requested, traffic numbers, tag numbers, etc. These
             data are minimally required to confirm the authenticity of the sample and its
             data.

       •      The chain-of-custody must be maintained at all times. The custody transfers
            . between different parties must be signed and dated.

3.5.2 Review Items.

A copy of the chain-of-custody form originated in the field and that returned from the
laboratory with the data are essential to confirm the identity of  the samples.  In addition,

-------
                                                         Procedure No.:  MI-PEST
                                                                       Revision:  1
                                                                 Date:  06/30/1995
                                                                     Page 20 of 20


the Region ni Shipping Record is essential to identify the field QC samples.  The chain-of-,
custody and  Shipping Record are generally located in front of the data package.

3.5.3  Evaluation Procedure

Ensure that the chain-of-custody form was signed and dated by the samplers, and a time
and date were entered for sample collection.  The laboratory copy of the cham-of-custody
must have the signature of the laboratory sample custodian.' Any errors on the form should
have been crossed out with a  single line through the  entry.  Verify mat all collected
samples have unique station identification,  traffic  numbers and sample  tag numbers.
Ensure that the Region m Shipping Record correctly reflects the information on the chain-
of-custody.                                                            .     •:..'.•

3.5.4  Action                                                 :
The action to be taken  in qualifying the data is highly dependent on the nature of the
problem:  Some errors  in paperwork are practically unavoidable in real situations.  An
effort should be made to reconcile the differences by cross checking.the field notebooks
a ;ainst the sampling paperwork.  Occasionally, the samplers may forget to sign the chain-
ojf-custody;  however, the  field notebooks  may  amply  describe  the  sampling event.
Problems are also inevitable in noting or cross-referencing sample tag numbers and traffic
numbers.  Generally, there are several alternate sources of information to .substantiate or
refute the problem.  ,                                         ,

3.5.5 Reporting

Any discrepancies found in the paperwork must be immediately brought to the attention of
the EPA RPM or SPCX  Clearly define the problems in a memorandum to the responsible
parties.  Attach marked copies of the chain-of-custody forms to substantiate the findings.

-------
                                                        Procedure No.:  M2-PEST
                                                                     Revision:  1
                                                               Date:  06/30/J995
                                                                    Pagel of 31-
                                 Appendix C-2
         Validation  of Pesticide/PCB Organic Analyte Data
                             Manual Level M2
                        1.  Purpose and Applicability

This procedure provides step-by-step instructions to manually validate the pesticide
organic data using the manual innovative data validation approach at Level M2.
approach focuses on the use of information contained on the CLP forms and a revie\jv of
chromatograms as summarized in Table M2-PEST-1.
•••' "''.:•'•   '.-..   "  "'-'"^'"'i-. ">ฃ'.  •.''•• ^'   *-'.•.•''••    -   '  '  .-  V •  ' .'•
The procedure is applicable to me Pesticide/PCB obtained using the Contract
Program Statement of Work (CLP SOW).  Hard copy data conforming to the applicable
CLP, SOW specifications are essential in orderto •carry out the procedure.

Data validated using this procedure are considered usable for  the following typ
purposes; however, the  data  users: must decide on a  case-by-case basis whether! the
procedure is suitable for their intended data uses. The suggested data uses are:  .= •   \
      •  " •     . •      •"          •.               •                     '  "  .•'     i'
       •     Oversight of activities led by other parties

       •     Comparison to action levels

       •     Initial site investigation

       •     Contamination sources

       •     Nature and extent of contamination
                   •    *               •
       •     Preliminary risk  assessment

       •     Risk assessment  with known high levels of toxics

       •     Feasibility study

       •     Preliminary design

       •     Treatability study

       •     Initial cleanup verification

-------
                                                         Procedure No.:  M2-PpST
                                                                      Revision:  1
                                                                   Date: 06/30/95
                                                                      Page 2 of 31"
                                Table M2-PEST-1
                       QC CHECKLIST FOR LEVEL M2  	
           CLP RAS PESTICIDES-POLYCHLORINATED BIPHENYLS
              QC MEASURES  .
. Action Level Notification
 GC/ECD Performance Check  -!
 Initial Calibration (CF)
 Initial Calibration (96RSD)
 Continuing Calibration (RPD)
.Laboratory Blank
 MS/MSD (%R, RPD)
 Field Quality Control (dup., blnk., PE samp.)

 Sample Paperwork  .      •      .

 Holding Time           •
 Retention Time
 Surrogate Recovery
 Dilution Factor

 Moisture Content

 Pesticide Cleanup Checks

 Chromatograms
 Raw Data

-------
                                                        Procedure No.;  M2-PEST
                                                                     Revision:  1
                                                               Date: 06/30/1995
                                                                    Page 3 of 31
                   2.  Quality Control Measures Checked

Table M2-PEST-1 highlights the quality control (QC) indicators evaluated under this data
validation procedure.
                                3.  Frocedure
                                ; '"'- • -, ' :- '-.'••  i ;"  -
The following subsections describe for each of the QC indicators the acceptance criteria,
location and retrieval of QC data, evaluation of the QC date, actions taken in the event the
QC acceptance  criteria are exceeded, and documentation of the QC  violations in a
standardized report form.

The pesticides data requirements to be checked are listed below:

       3.1    Action Level Notification

       3.2    Technical Holding Times (CCS—Contractual holding times only)
               I         •  .       •    •       •       -••,'••'•.
       3.3    GC/ECD Instrument Performance Check

       3.4    Initial  Calibration (CCS)

       3.5    Continuing Calibration (CCS)

       3.6    Blanks

       3.7    Surrogate Spikes (CCS)

       3.8    Matrix Spikes/Matrix Spike Duplicates

       3.9    Pesticide Cleanup Checks -

       3.10   Reported Contract Required Quantitation Limits (CRQLs)

Two forms have  been  developed  to  assist in the  performance and documentation of
implementing  Level  M2.  The first form,  M2-PEST-QUAL,  summarizes holding  time,
calibrations, blanks, surrogates, and internal standards.  The second form, M2-PEST-SPK,
summarizes surrogate and matrix spike quality control checks.  These forms are presented
on the following pages as Tables M2-PEST-2 and M2-PEST-3.

Reporting requirements for Level M2 are as follows:

-------
                                                          Procedure No.:  M2-PEST
                                                                       Revision:  1
                                                                 Date:  06/30/1995
                                                                      Page 4 of 31
       •     Hand annotate the Form I's, including:
             —    .Data validation qualifiers
             -     Sample identification number
             -     Sampling location

       •     Provide a narrative that includes:
             -  -   A statement that defines the level of the data review, i.e., M2
             —     Major and minor problems associated with the analysis
             -     Highlight issues that may have affected detection limits

       •     Include the following attachments:                         :
             -     Ust of data validation qualifiers
             - •    Support documentation including forms that support assigning data
                 •.' qualifiers^;,-.-;   •.'-.'•      •  '-.       ;  : :  .. . /   '.   ..
             -     Chain of custody form                         -
             -  >   Samples affected by calibration  should be listed on the appropriate
                    calibration^forms   .v      .      ••      .   .       -

Data qualifiers assigned in this review are as follows:                 ,
                            *''  .       '
Codes Relating  to  Identification (Confidence Concerning  Presence or  Absence  of
Compounds):

U        =    Not detected.   The associated number indicates approximate  sample
                concentration necessary to be detected.

(NO CODE) =  Confirmed identification.

B    .    =    Not detected substantially above the  level reported in laboratory or field
                blanks.

R        =    Unreliable result.  Analyte may or  may not be present in the sample.
                Supporting data necessary to confirm result.

N        =    Tentative identification.  Consider  present.  Special methods may  be
                needed to confirm its presence  or absence in future sampling efforts.

Codes Related  to  Quantitation  (can be used for   both positive  results  and  sample
quantitation limits):

J         =    Analyte present.   Reported value may not be accurate or precise.

-------
                      Method 608/8080
                      Cซซe No:	
                      SDG No:
                                                   Procedure No.:j M2-PEST
                                                             Revision: 1
                                                         Date:
                                       06/30/1995
                                      Rage 5 of 31
                                             TABLE M2-PEST-2
      PESTICIDE/PCBS QUALIFIER SUMMARY (CALIBRATIONS, BLANKS, HOLDING TIME, SURROGATES, INTERNAL
                                               STANDARDS)   .                                       f
Date Analyzed:
Matrix
• Instrument' ID:
Method Blank ID:
Extraction Date:
dilnmM; • Primary
• Secondary
Compound.
. Sample Identifier:


• . .- '

• '• •;.:;•;;•:'•• •' •• • •'.- •'
• .. •'.- :•-••: •'••;•':.••••.•'••••. ;• •
'. ''••"' -'-'-:v • . -'•'.*" '•• '•••
'y-. • / ;. ..- •- •

•..'..;'.•••'' • ••
Hold Time
Out. davs
Ext.




•





Anal.










Last IND Standard before Analysis
1










'• •,':•••. ••" -.'. -,. .' : • '
2



.. ••

• .•' *.V.
- . •• .'.
'•••'*.'



3



•-. •.' .'iซ.
•:•'•/•';??
••VN?$
•=•^^8
'^sM


. 4 ..



AV.;;:,>'
'$$&#ฃ
:5ft#M
^^>!
l>ป_ซi*'lr,^'ป
ซ^?l-''V- .;- •
;v.-..---v-
•*•?• ::
5: ซ
t
| .
1

.<•.....
?:- '
i* '.

,

,_.:.:.s.:;lvj,.!.i..;:;, •'
•..'••••'''. '• •. •' CaKbmrinnซt ' • ••. '• ' •.•••;. ••'.*ป!
InhVGaL -
ปRSD>10
'/' . ••'.', i ConLCal..d
aloha-BHC
beta-BHC
delta-BHC ' '
nmma-BHC fTJndanet • • '
Hentachlor
Aldrin
Hentachlor enoxide '
Endosulfan T
Dieldrin
4.4'-DDE
Endrin
Endosulfan II
4.4'-DDD
Endosulfan sulfate
4.4--DDT
Methoxvchlor
' Endrin ketone
alnha-chlordane
eamma-chlordane
Toxanhene
Aroclor-1016 fPCB-1016)
Aroclor-1221 fPCB-122H
Aroclor-1232 (PCB-12321
Aroclor-1242 fPCB-12421
Aroclor-1 248 fPCB- 1248)
Aroclor-1254(PCB-l254)
'Aroclor-1260 (PCB-1260)
Dinntvlchlorendnte (DBO SUIT.
ate/month
Tinie

.:• ••'••




,'





















,^ฃ:.'. -, C*u&
'•;*>?$•'•'•••"


-.; .,•....•



























"'T "
~. ;.: .• • •


'" ' " •


























iimklf^l
•J. '




'

























D> 15% or 20% .'.'••"•.•.•4:.-V:rt
" 4 '






























'.'•"•* ••.•'•
• •; '•.."








*




















'•'^t$
:••:?*??$
:r^;.^4

'.- ' '" ••/
. • ••.ป.'
* . • -u


.
•~->,
. ;.



















S;
•'Tr^v -'""' v-
ri?*w^*-ป^**''*1
?,:.-,;-•' -•

^.!- ;=•.-:-•.• •
..';•'> :' '


-- • • . .





















XI















uklifien


x f
**
\


!
|


i


















Validation Criteria:
Delected compounds
Undetected compounds
Primary Column
  %D < 15
  56D < 20
and
or
Secondary Column
   56D < 20
   ซD < 20

-------
                        Method 608/8080


                        Case No:	
                        SDGNo:
Procedure No.:  M2-PEST

           Revision: I
      Date: 06/30/1995
          Page 6 of 31
                          Table M2-PEST-3
 PESTICIDE/PCBS SURROGATE AND MATRIX SPIKE QUALITY CONTROL
                            SUMMARY

                       Surrogate Data Summary
Sample Identifier:
(Acceptance Range, %K):


'.••••'• .-'•-. '.'.' '
i •

i • ... "•' ' •'• •
!•-"'•'• • % ' - ;"••
.'.'.' ' '•' .' • "
, • '.'•..''•'•
' . ' ' • . •
'
'•'• '• ' ''• "'."••

Surrogate Recoveries, 5CR
Aqueous
(24-154)

• • ' -. • •'- .' .


,• '• • '• :•

•' * ";•'.-':•
..' ': .."'•ii.!;-
•• •••.•-•-.-;;•
' . ''. ' '
;. • ••• '•..;•.


Soil
(20450)
•' '• -•
•,. .-^ ->ป.-••• '•
'-;•':•;."•••;••.• ''
' \'
,••'• - ' •:••-..
' • •••• '':C'-'
•••'. ::''i^v^r'
••'•• •:&:••
-: :--:-;\ v". ':
'- ''""'-
•'••'.'•':'•• '
. -V •

Qualifier! (+/-)

. .'


, - ;. - . .
• - •
. V "'. ~~ '
. . ' '

. . •" • ' '



Surrogate — Dibutylchlorendate (DBC)
                                             . MS/MSP Data Summary
Spike Compound:
Matrix Spike, .Recovery,
%R
Range
Actual
Matrix Spike Duplicate,
, Recovery. %R
Range
Actual
MS/MSD
Precision. RPD
Range
Actual
Qualifiers
(+/-)
AQUEOUS SAMPLES ,
(jamma-BHC (Lindane)
Heptachlor
Aldrin
Dieldrin
Endrin
4.4'-DDT
56-123
40-131
40-120
52-126
56-121
, 38-127

v




56-123
40-131
40-120
52-126
56-121
38-127



,


14
20
22
18
21
27












SOIL SAMPLES •
panima-BHC (Lindane)
Heptachlor
Aldrin
Dieldrin
Endrin
4.4'-DDT
46-127
35-130
34-132
31-134
42-139
23-134






46-127
35-130
34-132
31-134
42-139
23-134






50
31
43
38
45
50







I

-



-------
                                                         Procedure No.: M2-PEST
                                                                      Revision:  1
                                                                 Date:  06/30/1995
                                                                     Page 7 of 31
K        =    Analyte present.  Reported value may be biased high.  Actual value is
                expected lower.

L        =    Analyte present.  Reported value may be biased low.  Actual value is
                expected to be higher.

UJ       =    Not detected, quantitation limit may be inaccurate or imprecise.

TJL       =    Not detected, quantitation limit is probably higher.

Other Codes                 :

Q    .    =    No analytical result    .  •   *  - .

*         =    Results reported from diluted analysis.
                     '     "     >''".*.
3.1  Action Level Notification

The purpose behind action level notification is to make the EPA Remedial Project Officer
(RPM) or the Site  Project Officer (SPO) aware of the potential human health risk at the
site.  In accordance with the Region ffl Hazardous Waste Division policy, the EPA RPM
or SPO must be promptly notified of any contaminant exceeding the established action level
or the 10-day health advisory limit.  The data for contaminants exceeding die action levels
must be validated as a top priority and reported to the RPM or SPQ.  Validation of the rest
of the data may then be completed normally.

3.1.1  Acceptance Criteria for Action Level Notification

EPA's Office of Solid Waste and Emergency Response has established 10-day advisory
limits or the  action levels for several organic compounds and elements of special health risk
concerns  based  on  the Safe Drinking Water Act.  The  pesticide organic  compounds and
their 10-day  health  advisory limits apply only to aqueous samples and are listed in  Table
M2-PEST-4. The criteria for action level notification are. as follows:

      •     The contaminant concentration must be equal to or above the established 10-
            day health advisory limits.

      •     Data for contaminants exceeding the action levels must be validated as a top
            priority.

      •     The  following  EPA  personnel  must be  notified of  the  action  level
            exceedances:

-------
                                                        Procedure No.:  M2-PEST
                                                                     Revision:  1
                                                                Date:  06/30/1995
                                                                    Page 8 of 31
                   EPA RPM or SPO

                   EPA Section Chiefs:
                          Site Investigation (SI)
                          Remedial
                          Enforcement
                          RCRA

                   EPA Section lexicologists:
                   -      Enforcement
                          Superfund
                          RCRA
                            .-.(••
                 •  '  '    * '   '-  .    ' •      -*" " -
      •      The remaining data validation.should be completed per normal procedures.

      •      Any special  instructions from the Hazardous Waste Division should be
             followed.

      •      Records should be'kept of the data review, action level notification and any
             follow up instructions and actions.
Table M2-PEST-4
PESTICIDES AND POLYCHLORINATED BIPHENYLS AND ACTION
LEVELS
Compound
ii
Chlordane
Heptachlor
Methoxychlor
Action
Level*
63
10
2,000
Compound
Endrin
Lindane
Toxaphene
Action
Level*
5
1,200
80
*A11 units are ug/1.
3.1.2  Data Requirements and Retrieval of Data

All data required to perform Level M2 validation, as detailed in the following sections, are
necessary for carrying out action level notification.  The.location of the data and their
retrieval procedures are also discussed below.

-------
                                                         Procedure No.; M2-PEST
                                                                      Revision: 1
                                                                 Date:  06/30/1995
                                                                     Page 9 of 31
3.1.3 Evaluation Procedure
The evaluation  process  preceding  action level  notification will primarily  consist of
comparing the results on Form I's with the action levels presented in Table M2-PEST-4.
Following the identification of the contaminants exceeding the action levels, focused data
validation should be .performed using the criteria,  and procedures described in the
appropriate sections below.

3.1.4 Action

The action resulting from focused data validation will be the notification of action level
exceedance to the personnel identified above in Section 3.1.1.
                   • •   '     ''     •'        •*• * . •     '. ' •
3.1.5 Reporting                                  '•'•'.

Copies of Form I's can be used to highlight the contaminants above the action levels.  The
findings of the focused validation can be summarized in  a memorandum/.land  the  data"
qualifiers resulting from focused validation may be written on the Form I*s. < The marked
up  forms should be clarified, that they represent validation of  only the contaminants
exceeding the action levels,  and not all data.

3.2 Technical Holding Times

The objective is to ascertain the validity of results based on the holding time of the sample
from time of collection  to-time of sample extraction and analysis.

3.2.1  Acceptance Criteria

Technical requirements  for sample holding times have only been established for water
matrices.  The holding times for soils are currently under investigation. When the results
are available they will be incorporated into the data evaluation process.  In addition, results
of holding time studies will  be incorporated into the data review criteria as the studies are
conducted and approved.  The maximum holding time, as stated in the current 40 CFR  Part
136, for pesticides and  Aroclbrs in cooled (@ 4ฐC) water samples is 7 days from sample
collection to extraction and 40 days from sample extraction to analysis. It is recommended
that pesticides and Aroclors in soil samples be extracted within 7 days of sample collection.

The contractual holding times, which differ from the technical holding times, state  that
extraction of water samples by separatory funnel must be completed within 5 days of
validated time of sample receipt (VTSR), extraction of water samples by continuous liquid-
liquid extraction procedures must be started within 5 days of VTSR, and soil/sediment
samples are to be extracted within 10 days  of VTSR. Also, contractually both water and

-------
                                                         Procedure No.: M2-PEST
                                                                      Revision:  1
                                                                 Date: 06/30/1995
                                                                     Page 10 of 31


soil  sample extracts must  be analyzed within  40 days of extraction.   However,  the
contractual delivery due date is either 14 days or 35 days after receipt in the laboratory of
the last sample in the SDG, depending on the contract.

3.2.2 Review Items

Form I PEST, EPA Sample Traffic Report, and/or chain-of-custody.

3.2.3 Evaluation Procedure

Technical holding times for sample extraction are established  by comparing the sample
collection date on the EPA Sample Traffic Report with the dates of extraction on Form  I
PEST. To determine if the samples were analyzed-within the holding time after extraction,
compare the dates of extraction on Form I PEST.             :   .        :

Verify that the traffic report indicates that the samples were received intact and iced.. If the
samples were not iced  or there were any problems with the samples upon receipt, then.
discrepancies in the sample condition jcould affect the data.       v       .

3.2.4 Action             -\(                            /

1.     If technical holding times are exceeded, qualify all detected compound results as
       estimated "J" and sample quantitation limits as estimated "UJ," except for PCBs
       which are not expected to degrade significantly during storage.

2.     If technical Holding times are grossly exceeded, either on the first analysis or upon
       reanalysis, the reviewer must use professional judgment to determine the reliability
       of the data and the effect of additional storage on the sample results.  The reviewer
       may determine that detected compound results or the associated quantitation limits
       are  approximates and should be qualified with "J"  or  "UJ," respectively.  The
       reviewer may determine that nondetected target compound data are unusable (R).

3.     Due to limited information concerning holding times for soil samples, water holding
       time criteria should be applied.

4.     The reviewer should  also be aware of the scenario in  which the laboratory has
       exceeded the technical holding times, but met contractual holding times.  In this
       case, the data reviewer should  notify  the Regional TPO  (where  samples  were
       collected)  and/or RSCC that shipment  delays  have occurred  so  that the field
       problem can be corrected.   The reviewer may  pass this information on to  the
       laboratory's TPO, but  should explain  that contractually the  laboratory  met  the
       requirements.

-------
                                                       Procedure No.: M2-PEST
                                                                    Revision: 1
                                                               Date: 06/30/1995
                                                                   Page 11 of 31
3.3 GC/ECD Instrument Performance Check

Performance checks on the gas chromatpgraph with electron capture detector, (GC/ECD)
system are performed to  ensure adequate resolution and instrument sensitivity.  These
criteria are not sample specific.  Conformance is determined using standard materials;
therefore, these criteria should be met in all circumstances.

3.3.1  Acceptance Criteria

1.    Resolution Check Mixture

      a.     The Resolution Check Mixture must be analyzed at the beginning of every
             initial calibration  sequence, on eath'GC column, and instrument used for
             analysis.  The Resolution Check Mixture contains the following pesticides
             and surrogates: ^:                                      '    ..'-•'

                   gamma-Chlordane
                   EndosulfanI
                   4,4'-DDE
                   Dieldriri
                   Endosulfan sulfate
                   Endrinketone
                   Methoxychlor
                   Tetrachlorb-m-xylene
                   DecachlorobiphenyL.,

      b.     The depth of the valley between two adjacent peaks in the Resolution Check
             Mixture must be greater than or equal to 60.0 percent of the height of the
             shorter peak.                                                     ~

2.     Performance Evaluation Mixtures

      a.     The Performance Evaluation  Mixture (PEM) must be  analyzed at the
             beginning (following the resolution  check  mixture) and at the end of the
             initial  calibration  sequence.   The  PEM  must  also be  analyzed at the
             beginning of every other 12-hour analytical period.  The PEM contains the
             following pesticides and surrogates.

             •     gamma-BHC
             •     alpha-BHC
             •     4,4'-DDT
             •     beta-BHC

-------
                                                       Procedure No.:  M2-PEST
                                                                    Revision:  1
                                                               Date:  06/30/1995
                                                                  Page 12 of 31
             •     Endrin
             •     Methoxychlor
             •     Tetrachloro-m-xylene
             •     Decachlorobiphenyl
      b.     The resolution of adjacent peaks for the PEM injections in each calibration
             (initial and continuing) must be 100 percent for both GC columns.  •
                      .•               -                  •   ••         -     •
      c.     The absolute retention times of each of the single component pesticides and
             surrogates in all PEM analyses must be within the specific^retention time
             windows centered  around the mean retention times determined from the
             three-point initial calibration using the individual Standard Mixtures. , A list
             of the retention time windows is found in Attachment 1, Table &ฃ '^
            • ".  •"••:•/•'  -.'...-    ,-  • •      .     '••"   '-,- •<:&ฃ$$ฃ•$* •'•"'.;.
             For example,  for a given  pesticide  the mean, retention  time  is first
             determined from the initial calibration and found to be 12.69 minutes.  The
             retention time window for this pesticide is ฑ0.05 minutes, ; Therefore, the
             calculated retention time window would range from 12.64 to 12.74 minutes.

      d.     The relative percent difference (RPD) between the calculated amount and the
             true amount for each of the single component pesticides and surrogates in
             the PEM analyses must be less than or equal to 25.0 percent. _
                  ij                                   .               •
      e.     The percent breakdown is the amount of decomposition that 4,4'-DDT and
             Endrin undergo when analyzed on the GC column.  For Endrin, the percent
             breakdown is determined by the presence of Endrin aldehyde and/or Endrin
             ketone in the GC chromatogram. For 4,4'-DDT, the percent breakdown is
             determined from the presence of 4,4'-DDD and/or 4,4'-DDE in the GC
             chromatogram. The equations used to verify these calculations are provided
             in Attachment 1, Equations A and B.

             i.     The percent breakdown for both 4,4'-DDT and Endrin in each PEM
                   must be less than or equal to 20.0 percent for both GC columns.

             ii.    The combined percent breakdown for 4,4'-DDT and Endrin in each
                   PEM  must  be less than or equal to 30.0 percent  for both GC
                   columns.

3.3.2  Review Items

Form VI PEST-4,  Form VH PEST-1 Form VIII PEST, and chromatograms.

-------
                                                        Procedure No.: M2-PEST
                                                                     Revision:  1
                                                                Date: 06/30/1995
                                                                    Page 13 of 31
3.3.3 Evaluation Procedure

1.     Resolution Check Mixture
                              '
             Verify from the Form VIE PEST that  the resolution check mixture was
             analyzed  ai the beginning of the initial  calibration sequence on -each GC
             cx)lunm and instrument used for analysis.
                 •                     "                        '
      b.   ,  Check the resolution check mixture data and Form VI PEST-4 to verify that
             the resolution criterion  between two adjacent  peaks for  the  required
             compounds is less'than or equal to 60 percent.
           '   '•'                    "                               '   '
2.     Performance Evaluation Mixture
      Verify from ^brii^EST that the Performance Evaluation Mixture (PEM)
      was analyzed at^^po^erl^rrequency and position sequence.
               :'  '-y;-fi^^^^^':'-"• - •                    •'••'•.•'
      a.     Check the PEM data from the initial and continuing calibrations to verify
             that the'resolution between adjacent peaks  is 100 percent on both GC
             columns.  '':  ;"v:                             •"•     v

      b.     Check the PEM data from the initial and continuing calibrations and Form
             VH PEST-1 to verify that the absolute retention times for the pesticides in
             each analysis are within the calculated retention time windows based on the
             mean  RT from  the three-point  initial calibration using  equations  and
             examples found in Attachment 1, Table 1.                         "   ••

      d.     Verify that the relative percent difference (RPD) between the calculated
             amount and the true amount for each of the pesticides and surrogates is less
             than or equal to 25.0 percent.  '

      e.     Verify that the individual breakdowns for 4,4'-DDT and Endrin are less  than
             or equal to 20.0 percent, and that the combined breakdown is less  than or
             equal to 30.0 percent.
3.3.4  Action
      Resolution Check Mixture: If resolution criteria are not met the quantitative results
      may not be accurate due to inadequate resolution.  Detected target compounds that
      were not  adequately  resolved  should be  qualified  with  "J."   Qualitative
      identifications may also  be  questionable if  coelution  exists.   Nondetects with
      retention times in the region of coelution may not be valid, depending on the extent

-------
                                                         Procedure No.:  M2-PEST
                                                                       Revision:  1
                                                                 Date:  06/30/1995
                                                                     Page 14 of 31
       of the problem.  Professional judgment should be used to determine the need to
       qualify data as unusable (R).

2.     Performance Evaluation Mixture Retention Times:  Retention time windows are
       used in qualitative identification. If the retention times of the pesticides in the PEM
       do not fall within the retention time windows, the associated sample results should
       be carefully evaluated.  All samples injected after the last in-control standard are
       potentially affected.  It should be noted for TPO action if the PEM retention time
       criteria are grossly exceeded.                                      ~  •
                         • • - - ;    -.. •       .'            '     '       -is .-.;••
       a.     For the affected samples, check to see if the sample chromatograms contain
             any peaks that are close to  the  expected retention time window of the.
             pesticide of interest.  If no peaks are* present either within,or .close .to. the
             retention time window  of the deviant target pesticide compound, then there
             is usually no effect on  the data (i.e., nondetected values can be considered
             valid). Sample data that are potentially affected by standards not meeting
         .    the retention time windows should be noted in the data review'narrative.

       b.     If  the affected sample chromatograms contain peaks /which -may  be  of
             concern (i.e., above the CRQL and either close to or within the expected
             retention time window  of the analyte of interest), then the reviewer should
             determine the extent of the effect on the data  and may  choose to qualify
             detected target compounds "N" and nondetected target compounds "UJ."
             In  some cases, additional effort by the reviewer may  be necessary  to
             determine if sample peaks represent the compounds of interest, for example:

             i.            The reviewer can examine the  data package for the presence
                          of three or more standards containing the pesticide of interest
                          that  were run  within a-  72-hour period during which the
                          sample was analyzed.

             ii.            If three  or more such standards  are present, the mean and
                          standard  deviation  of  the  retention time  window  can  be
                          reevaluated.

             Hi.           If all standards  and matrix spikes fall within the revised
                          window,  the valid positive or negative sample results can  be
                          determined using this window.

             iv.           The narrative should identify the additional efforts  taken  by
                          the reviewer and the resultant impact on data usability.  In
                          addition,  the  support documentation   should  contain  all
                          calculations and comparisons generated by the reviewer.

-------
                                                         Procedure No.:  M2-PEST
                                                                      Revision:  1
                                                                Date: 06/30/1995
                                                                    Page 15 of 31


       c.     If the reviewer cannot do anything with the data to resolve the problem of
             concern, all positive results and quantitation limits should be qualified "R."

3.     Performance Evaluation Mixture Resolution: If PEM resolution criteria are not met
       then the quantitative results may not be accurate  due to  inadequate resolution.
       Positive sample results  for compounds that were not adequately resolved should be
       qualified with "J."  Qualitative identifications may be questionable if coelution
      .exists.  Nohdetected target compounds that elute in the region of the coelution may
       not be valid depending on the extent of the coelution problem,   Professional
      judgment should be tosed td qualify data as unusable (R).

4      If RPD criteria are not met/ qualify all associated positive results  generated during
       the analytical sequence  with-v'J" and thejample quantitation limits for nondetected
       target compounds wito  "OT."  ;-V       '
                  .      ;•';:  :4K-;- '''•:.'•.•.  ••;     •"•.'.•'.•'•':'•.'.
5.     4,4'-DDT/Endrin Breakdowii:
       a.     If 4,4'-DDT breakdown is greater than 20.0 percent:

             i.     Qualify all positive results for DDT as "L" biased low.  If DDT
                   was not detected, but DDD and DDE are detected, then qualify the
                   quantitation limit for DDT as unusable (R).

             ii.    Qualify positive results for  DDD and/or  DDE as  presumptively
                   present at an estimated quantity (N).

       b.     If Endrin breakdown is greater than 20.0 percent:

             i.     Qualify all positive results for Endrin  with "J."  If Endrin was not
                   detected,  but Endrin aldehyde and Endrin ketone are detected,  then
                   qualify the quantitation limit for Endrin as unusable (R).

             ii.    Qualify positive results for Endrin ketone as presumptively present at
                   an approximated quantity (N).

      c.     If the combined 4,4'-DDT and Endrin breakdown is  greater than 30.0
             percent:

             i.     Qualify all positive results for DDT and Endrin with "J."  If Endrin
                   was not  detected,  but  Endrin  aldehyde and Endrin ketone  are
                   detected,  then qualify the quantitation limit for Endrin as  unusable
                   (R). If DDT was not detected, but DDD and DDE are detected, then
                   qualify the quantitation limit for DDT as unusable (R).

-------
                                                         Procedure No.:  M2-PEST
                                                                       Revision:  1
                                                                 Date: 06/30/1995
                                                                     Page 16 of 31
             ii.     Qualify positive results for Endrin ketone as presumptively present at
                    an estimated quantity (N).  Qualify positive results for DDD and/or
                    DDE as presumptively present at an approximated quantity (N).
                    '.''.•      '               i                 '
6.     Potential effects on the  sample data resulting from the initial calibration problems
    /  should be noted in the data review narrative.

3.4 Initial Calibrkiim H    >

Compliance requirements for satisfactory initial calibration are established to ensure that
the instrument is capable of producing acceptable qualitative and quantitative data for
pesticide and  Aroclor  target  compounds.   Initial  calibration  demonstrates  that  the
instrument is capable of acceptable performance at the beginning of the analytical sequence
and of producing a linear calibration curve.                                .

3.4.1  Acceptance Criteria .

\.     Individual Standard Mixtures

       a.     Individual  Standard  Mixtures A  and  B (containing  all of the single
             component pesticides and surrogates) must be analyzed at low, midpoint,
             and high  levels  during the  initial calibration, on each GC  column and
             instrument used for analysis.

       b.     The resolution between  any two adjacent peaks  in the midpoint concentration
             of Individual Standard  Mixtures A  and B in the initial calibration must be
             greater than or equal to .90.0 percent on both columns.

       c.     The absolute retention times of each of the single component pesticides and
             surrogates  are determined from three-point  initial calibration  using  the
             Individual Standard Mixtures.   A  list of .the  retention time  windows  are
             included in Attachment  1, Table 1.              ,

       d.     At least one chromatogram from each of the Individual Standard Mixtures A
             and B must yield  peaks that  give recorder  deflections  between 50  to
             100 percent of full scale.

       e.     The concentrations of the low, medium, and high level standards containing
             all of the  single  component  pesticides and surrogates (Individual Standard
             Mixtures A and B) must meet the following criteria on both GC columns.

             The low point corresponds to the CRQL for  each analyte.  The midpoint
             concentration must be four times the low point. The high point must be at
             least 16 times the low point, but a higher concentration  may be chosen.

-------
                                                        Procedure No.:  M2-PEST
                                                                     Revision:  1
                                                               Date:  06/30/1995
                                                                   Page 17 of 31


       f.     The Percent Relative Standard Deviation (%RSD) of the calibration factors
             for eaqh of the single component pesticides and surrogates in the initial
             calibration on both columns for-Individual Standard Mixtures A and B must
             be less than or equal to 20.0 percent, except as noted below.  For the two
             surrogates, the %RSD must be less than or equal to 30.0 percent.  Up to
             two, single component target pesticides (other  than the. surrogates) per
             column may exceed the 20.0 percent limit but the %RSD must be less than
             or equal  to  30.0 percent.   Calculation is included in   Attachment  1,
             Equation D.  .•   ''      '   •      ./•.-   '  -\ ;^v-r;'\.-;  •

       Note:  Either peak area or peak height may be used to calculate the calibration
             factors that are, in  turn, used to calculate  %RSD. However, the type of
             peak measurement  used to calculate/each calibration factor  for a given
             compound must be consistent.  For example, if peak area is used to calculate
             the low point calibration factor for  endrin, then the^mid-^anci high-point
             calibration factors for endrin must also be calculated using peak area.

2.     Multi-Component Target Compounds

       a.     The multi-component target compounds (the 7 Aroclor and Toxaphene) must
             each be analyzed separately  at a  single concentration level during the initial
             calibration sequence. The analysis of the multi-component target compounds
             must also contain the pesticide surrogates.

       b.     For each  multi-component analyte, the retention times are determined for
             three to five peaks.  A retention time window of ฑ0.07 minutes is used to
             determine retention time windows for all multi-component analyte peaks.

       c.     Calibration factor data must be determined for each peak selected from the
             multi-component analysis.

3.4.2  Review Items

Form VI PEST-1, 2, 3, and 4, Form VII PEST-1, "Form VHI PEST, and chromatograms.

3.4.3  Evaluation Procedures

1.     Individual Standard Mixtures

       a.     Verify from  the  Form VIII  PEST that the Individual Standard
             Mixtures A and B were analyzed at the proper frequency-on each
             GC column and instrument used for analysis.

-------
                                                        Procedure No.:  M2-PEST
                                                                    Revision:  1
                                                               Date: 06/30/1995
                                                                   Page 18 of 31
2.     Multi-Component Target Compounds

       a.     Verify  from  the  Form VIE PEST  that each of the  multi-
             component target  compounds were analyzed  at  the  required
             frequency.

       b.     Check the data for the multi-component target compounds  and
             Form PEST VI-3 to verify that a least three peaks were used for
             calibration and that retention time and calibration factor data are
             available for each peak.                           •--
3.4.4 Action

1.    If the initial calibration sequence was not followed as required, then professional
      judgment must be used to evaluate the effect of the non-compliance on the sample
      data.  If the requirements for the initial calibration sequence were not met, then this
      should be noted for TPO action.  If the non-compliance has a potential effect on the
      data, then the data should be qualified according to the professional judgment of the
      reviewer and this should be noted in the data review narrative.

2.    If resolution criteria are not met,  then the quantitative results may not be accurate
      due to peak overlap and lack of adequate resolution.  Positive sample results for
      compounds  that were not  adequately  resolved, should be  qualified  with "J."
      Qualitative  identifications may  be questionable if coelution  exists.  Nondetected
      target compounds that elute in the region of coelution may not be valid depending
      on the extent of the coelution problem.  Professional judgment should be used to
      qualify data as unusable (R).              •
3.    If the %RSD linearity criteria are not met for the compound(s) being quantified,
      qualify  all associated positive  quantitative  results  with "J" and  the  sample
      quantitation limits for nondetected target compounds with "UJ."

4.    Potential effects on the sample data due to problems with calibration should be
      noted in the data  review narrative.  If the data reviewer has knowledge that the
      laboratory has repeatedly failed  to comply with the requirements for frequency,
      linearity, retention time, or resolution, the data reviewer should notify the TPO.

3.5.  Continuing Calibration

Compliance requirements for satisfactory  instrument calibration are established to ensure
that the instrument is capable of producing acceptable qualitative and quantitative data.

-------
                                                        Procedure No.: M2-PEST
                                                                     Revision:  1
                                                               Date:  06/30/1995
                                                                   Page 19 of 31
Continuing calibration checks and documents satisfactory performance of the instrument
over specific time periods during sample analysis.  To verify the calibration and evaluate
instrument performance, continuing calibration is performed, consisting of the analyses of
instrument  blanks, the PEM,  and the midpoint  concentration of Individual Standard
Mixtures A and B.

3.5.1  Acceptance Criteria

1.     An instrument blank and the PEM must bracket one end of a 12-hour period during
       which  samples are analyzed, and a  second instrument blank and the midpoint
       concentration of Individual Standard Mixtures A and 8 must bracket the other end
       of the 12-hbur period.                                   -'r^S^;
              ''    - i -, '.':•••   :  •'.•   '         -jf- "'       •      .''. '"<•••:' '•;•?••'•Ashts-V-r '__
2.     The resolution between any two adjacent peaks in the midpoint'Concentration of
      .Individual  Standard  Mixtures  A and  B  must be greater rthan^ror equal to
       90.0 percent     ';•••_,-     •    .            •  • ;-.'••• ••'^"-ป4/.';'-  •„  '   -.

3.     The absolute retention time for each single component pesticide and surrogate in the
       midpoint concentration of Individual Standard Mixtures A and B must be within the
       retention time windows determined from the initial calibration.    ""•'

4.     The RPD between the calculated amount and the true  amounrfor each of the
       pesticides and surrogates in the midpoint concentration of the Individual Standard
       Mixtures A and B must not exceed 25.0 percent.            :

3.5.2  Review Items

Form Vn PEST-1 and 2, Form VHI PEST, and chromatograms.

3.5.3  Evaluation Procedure

1.     Check  the Form Vm PEST to verify that the instrument blanks, PEMs,  and
       Individual Standard Mixtures were analyzed at the proper  frequency and that no
       more than 12 hours  was  elapsed  between  continuing calibration brackets in an
       ongoing analytical sequence.

2.     Check the data for the midpoint concentration of Individual Standard Mixtures A
       and B to verify that the resolution between any two adjacent peaks is greater than or
       equal to 90.0 percent.

3.     Check the data for each of the single component pesticides and surrogates in the
       midpoint concentration of Individual Standard Mixtures  A and B and  Form  VII

-------
                                                         Procedure Nov! M2-PEST
                                                                      Revision:  1
                                                                 Date: 06/30/1995
                                                                     Page 20 of 31
       PEST-2 to verify  that the absolute retention  times are within the appropriate
       retention time windows.

4.     Check the data from the midpoint concentration  of Individual Standard Mixtures A
       and B and Form VH PEST-2 to verify that the RPD between the calculated amount
       and the true amount for each of the pesticides and surrogates is less than or equal to
       25.0 percent.

3.5.4 Action

1.     If  the  continuing  calibration  sequence  was  not  followed  as required,  then
       professional judgment must be used to evaluate the effect of the noncompliance on
       the sample data.  If the requirements for the'continuing calibration sequence were
       not met, then this should :1>e noted for TPO action.  If the noncompliance has a
      . potential effect on  the data, then the data should be qualified according to the
       professional judgment of me reviewer and this should be noted in the data review
       narrative.                V                      :       ,

2.     If resolution criteria are not met than the quantitative results may not be accurate
       due to inadequate resolution.  Positive sample results for compounds that were not
       adequately resolved should be qualified with " J." Qualitative identification may be
       questionable if coelution exists.   Nondetected target compounds  that  elute in the
       region of  coelution may not be  valid  depending on the extent  of the coelution
       problem.  Professional judgment should be used  to qualify data as  unusable (R).

3.     Retention time windows are used in qualitative identification.  If  the standards do
       not fall within the retention time windows, the associated sample results should be
       carefully evaluated.  All samples injected after the last in-control  standard are
       potentially effected.    ,

       a.     For the affected samples, check to see if  the sample chromatograms contain
             any peaks that are close  to  the expected retention time window  of the
             pesticide of interest.  If no peaks are present either within or close to the
             retention time window  of the  deviant  target pesticide compound,  then
             nondetected  values can be considered valid.  Sample data that is potentially
             affected by the standards not meeting the retention time windows should be
             noted in the  data review narrative. If the retention time  window criteria are
             grossly exceeded,  then this should be noted for TPO action.

       b.     If  the  affected sample chromatograms  contain  peaks  which  may be of
             concern (i.e., above the CRQL  and either close to or within  the expected
             retention time window of the pesticide of interest), then  the reviewer should

-------
                                                           Procedure No.:  M2-PEST
                                                                         Revision:  1
                                                                   Date:  06/30/1995
                                                                       Page 21 of 31-
              follow the guidelines provided in Section 3.3 to determine the extent of the
              effect on the data.

 4.     If the RPD is greater than 25 percent for the compound(s) being quantified, qualify
        all associated positive quantitative results with "J" and the sample quantitation
        iimits for nondetects with "UJ."                       .
•;.   '             '    . " ""•••>'' '•''..-•.                      '         " •'  '   '.'•' •  •' •
 5.     Potential effects  on the  sample data due  to  problems with calibration  should be
        noted in the data review narrative.  If the data reviewer has knowledge that  the
    '    laboratory has repeatedly failed to comply with the requirements for frequency,
        linearity, retention time, resolution, or DDT/Endrin breakdown, the  data reviewer
        should notify the TPO.

 3.6  Blanks

 The purpose of laboratory  (or field) blank analyses  is to determine the existence and
 magnitude of contamination  problems resulting from laboratory  (or field) activities. The-
 criteria for evaluation of laboratory blanks apply to any blank associated with the samples
 (e.g., method blanks, instrument blanks, and sulfur cleanup blanks).  If problems with any
 blank exist,  all associated data  must be carefully evaluated to determine whether or not
 there is an inherent variability in the data, or if the problem is an isolated occurrence not
 affecting other data.

 3.6.1  Acceptance Criteria

 1.    No contaminants should be present in the blanks.

 2.    Frequency:

       a.     Method Blanks—A method blank  analysis  must be  performed  for each
              20 samples of similar  matrix in each  sample delivery  group (SDG)  or
              whenever a sample extraction procedure  is performed.

       b.     instrument Blanks—An acceptable instrument blank must be run at least once-
              every  12 hours  and  immediately  prior  to the  analysis of  either  the
              performance evaluation mixture or Individual Standard Mixtures A  and B,
              depending on the place in the analysis  sequence.

       c.     Sulfur Cleanup Blanks—A sulfur cleanup blank must be analyzed whenever
              part of a set of samples extracted together requires sulfur cleanup.   If the
              entire set of samples associated with a method blank requires sulfur cleanup,
              then the  method blank also serves the  purpose of a  sulfur blank and  no
              separate sulfur blank is required.

-------
                                                         Procedure No.: M2-PEST
                                                                      Revision:  1
                                                               '  Date: 06/30/1995
                                                                     Page 22 of 31
3,6.2  Review Items

Form I PEST, Form IV PEST, and chromatograms.

3.6.3  Evaluation Procedure

1.     Review the results of all associated blanks, Form I PEST, Form IV PEST, and
       chromatograms to  evaluate the presence of TCL pesticides.

2.     Verify that method blank analysis  has been  reported per SDG, per matrix, per
       concentration level, for each GC system used  to analyze samples,-- and  for each
       extraction batch.                                                   v   v

3.     Verify that the method blank analyses do not contain  any target pesticide or
       Aroclor/Toxaphene at greater than its Contract  Required Quantitation Limits
       (CRQL).          :                                ,.              ;

4.     For the surrogates in each method blank, verify that .the observed retention times'
       are  within  the  appropriate  retention  time windows  calculated 'from  the initial
       calibration.                               ,.                     .

5.     Verify that the instrument blank analysis has been performed every; 12. hours as part
       of the continuing  calibration and following a sample  analysis which contains an
       analyte(s) at high concentration(s), and that the instrument blanks do not contain any
       target analytes above one-half the CRQL,  assuming that  the  material in the
       instrument resulted from the extraction of a 1-L  water sample.

6.     Verify that the sulfur cleanup blanks were  analyzed at the required frequency and
       that they do not contain any target compound above the CRQL, assuming that the
       material in the instrument resulted from the extraction of a 1-L water sample.  If a
       separate sulfur cleanup blank was prepared, one  version of Form IV PEST should
       be completed associating  all the 'samples  with  the method blank, and  a second
       version of Form IV PEST should be completed listing only those samples associated
       with the separate sulfur cleanup blank.

3.6.4 Action

If the appropriate blanks were not analyzed with the frequency described in Criteria 3, 4,
and 5,  then the data reviewer  should use  professional judgment to  determine if the
associated sample data should be qualified.  The reviewer may need to obtain additional
information from the laboratory.  The situation should  be brought to the attention of the
TPO.

-------
                                                          Procedure No.:  M2-PEST
                                                                        Revision:  1
                                                                  Date:  06/30/1995
                                                                      Page 23 of 31
Action in the case of unsuitable blank results depends on the circumstances and the origin
of the blank.   Detected compound results should be reported and qualified "B" if the
concentration of the compound in the sample is  less than or equal to 5 times  (5x) the
amount in the blank.  In instances where more than one blank is associated with a given
sample, qualification should be based on a comparison with the associated blank having the
highest concentration of a contaminant.  The results must not be corrected by subtracting
the blank value.    •   r--'--v"^-;''/:;-  •  :"  •   .   "•••            -v'-''•••;;:-.--;•'•.'•.
Specific actions are as follows:
1.     If a target pesticide or Aroclor/Toxaphene is found in the blank but hot found in the
       sample(s), no qualification is required.   If the contaminant(s) is  found at level(s)
       significantly greater than the CRQL, then-this should be noted in the data review
       narrative.         -      "                                    !
            '-         i      ป   **-/""         '                 - -,  .-'--:••.•"."._•.
              • . -           ป            •            '    '    .      '         .       '
2.     Any pesticide or  Aroclor/Toxaphene detected in the sample, that was also detected
       in any associated blank, is qualified if  the sample concentration  is less than  five.
       times (5x) the blank concentration.  The quantitation limit may also be elevated:
       Typically, the sample CRQL is elevated to the concentration found in the sample.
       The reviewer should use professional judgment to determine if further elevation of
       the CRQL is required.

       The reviewer should note that  analyte concentrations calculated for method blank
       may not involve  the same weights, volumes, or dilution factors as the associated
       samples.  These factors must be taken into consideration when applying the "Sx"
       criteria,  such that a comparison of the  total amount of contamination is actually
       made.

       In addition, there may be instances when little or no contamination was present in
       the associated blanks, but qualification of the sample, was deemed necessary.
       Contamination introduced  through dilution  is one example.  Although it is not
       always possible to. determine,  instances of this occurring  can be detected when
       contaminants are found  in the  diluted sample result, but  absent  in the undiluted
       sample result.  Since both results are not routinely reported, it may be possible to
       verify this source of contamination.  However, if the reviewer determines that the
       contamination is  from a source other than the sample, he/she should  qualify the
       data.  In this  case, the "5x"  rule  does not  apply; the sample  value  should be
       reported and qualified "B" and  a note should be added to the narrative.

3.     If gross contamination exists (i.e., saturated peaks), all affected compounds  in the
       associated samples should be qualified as unusable (R), due to interference.  This
       should be noted in the data review narrative if the contamination is suspected of
       having an effect on the sample  results.

-------
                                                          Procedure No.: M2-PEST
                                                                       Revision: 1
                                                                 Date:  06/30/1995
                                                                     Page 24 of 31
4.     If inordinate amounts of other target pesticides or Aroclors/Toxaphene are found at
       low levels in the blank(s), it may be indicative of a problem at the laboratory and
       should be noted in the data review narrative.

5.     If  an instrument blank was not analyzed following a sample analysis  which
       contained an analyte(s) athighconcปntration(s), sample analysis results after the
       high .concentration sample must be evaluated for carryover.  Professional judgment
       should be used to determine if instrument cross-contamination has effected any
                   . •  s\- '••-'.*      -                     ,. . ., X •'.''••         *
:       positive compound identificationฎ, and if so, detected compound results .should be
       qualified.  If instrument cross-contamination is suggested, then this should be noted
       in  the data review narrative if the cross-contamination is suspected of having an
       effect on the sample results.                                 ^:U->,
                      '     '                      '
The following  are  examples of applying the blank qualification guidelines.   Certain
circumstances may warrant deviations from these guidelines:       /v.  ;^j>
                                                             • ••• '-.^i*:ซ•''.'?"'-;• :•'.". '  ' •
                                                                •  ซ' •"" . .ซ   " -•
Example 1:'  Sample result is greater than the  CRQL, but is less than the 5x
             multiole of the blank result                            •
                    multiple of the blank result.
                          Blank Result            :         1.0
                          CRQL                           0.5
                          Sample Result                    4.0
                          Qualified Sample Result           4.0B

                    In  this case,  sample results  less than 5.0 (or 5 x 1.0) would be
                    qualified as nondetected target compounds.
      Example 2:   Sample result is less than the CRQL,  and is also less than the 5x
                    multiple of the blank result.

                                                           5jc

                          Blank Result                      1.0
                          CRQL                           0.5
                          Sample Result                    0.4J
                          Qualified Sample Result           0.4B

-------
                                                          Procedure No.: M2-PEST
                                                                       Revision: 1
                                                                  Date:  06/30/1995
                                                                      Page 25 of 31
       Example 3:  Sample result is greater than the 5x multiple of the blank result.
                           Blank Result                     1.0
                           CRQL                          0.5
                           Sample Result                   10.0
                         -Qualified Sample Result           10.0 v

                    In this case, the sample result exceeded the adjusted blank result (5x
                    1.0) and the sample result is not qualified.
                    . '.-•• ; -iv-Jfti^^^sSv; v v ••
' 3.7  'Surrogate
.   ".   •'    • •  ;• ".
 Laboratory performance oniindi>idual samples is established by means of spiking samples
 prior to extraction and analysis 'to determine surrogate spike recoveries.  All samples are
 spiked with  surrogate compounds prior to sample extraction.   Hie  evaluation of the
 recovery results of these surrogate .spikes is not: necessarily straightforward.  The sample
 itself may produce effects due to such factors as interferences and high concentrations of
 target and/or nontarget ahalytes.   Since the effects of the sample matrix are frequently
 outside the control of the laboratory and may present relatively unique problems, the
 evaluation and  review of data based on specific sample results is frequently, subjective and
 demands analytical experience and  professional judgment.   Accordingly,  this  section
 consists primarily of guidelines, in some cases with  several optional approaches suggested.

 3.7.1  Acceptance Criteria

 1.     Two surrogate spikes, tetrachloro-m-xylene  and  decachlorobiphenyl, are added to
       all  samples, Individual, Standard Mixtures,  PEMs,  blanks, and matrix spikes to
       measure their recovery in sample and blank matrices.

 2.     The advisory limits for recovery of the surrogates tetrachloro-m-xylene (TCX) and
       decachlorobiphenyl (DCB) are 60 to 150 percent  for both water and soil samples.

 3.     The retention times of both of the surrogates in the PEM,  Individual Standard
       Mixtures,  and samples must be within the calculated retention time windows. TCX
       must be within ฑ0.05 minutes, and DCB must be within  ฑ0.10 minutes of the
       mean retention time determined from the initial calibration.

3.7.2  Review Items

Form II PEST, Form Vffl PEST, and chromatograms.

-------
                                                         Procedure No^:  M2-PEST
                                                                      Revision:  1
                                                                 Date: 06/30/1995
                                                                     Page 26 of 31
3.7.3 Evaluation Procedures
1.     Verify that the recoveries on the Surrogate Recovery Form n PEST are accurate
       and within the advisory limits and  that the retention  times  on  the Pesticide
     ,  Analytical Sequence Form Vm PEST are accurate and within,the retention time
 •  "    limits.-'  . • .;  ''/ •"...''." •   . .          •         '  .' ,    -•.'-.-..

2.     If low surrogate •recoveries are observed, the reviewer should investigate whether
       me low recoveries were a result of sample dilution.

3.     In the special case of a'blank analysis  with surrogates out of  specification, the
       reviewer must give special consideration to the validity of associated sample data.
       The basic concern is whether the blank problems represent an isolated problem with
       the blank alone, or whether  there is  a  fundamental problem with the analytical
     .  process.   For  example, if one  or more samples in the batch show •acceptable
       surrogate recoveries, the reviewer may choose to consider the blank problem to be
       an isolated occurrence.                                       .

3.7.4  Action

1.     If surrogate spike recoveries are.outside of advisory limits, the following guidance
       is suggested.  Professional judgment must be used in applying these-criteria.

       a.     If low recoveries (i.e., between 10 and 60 percent) are obtained,f this may
             be an indication of a  low bias in sample  results  and associated detected
             compound data should  be qualified ' 'L'' and quantitation limits 4 'UL.''

       b.     If high recoveries (i.e, greater than 150 percent) are obtained, this  may be
             an indication of a high bias due to co-eluting interferences. Qualify
             associated detected compound data with "K", nondetected analytes do not
             require qualification.

       c.     If either pesticide surrogate recovery is reported as between zero percent and
             10 percent, the reviewer should examine the sample chromatogram to assess
             the qualitative validity  of the analysis.  If low surrogate recoveries are found
             to be due to sample dilution, then professional judgment should be used to
             determine if the resulting data should be qualified.  If sample dilution is not
             a  factor,  then detected  target compounds may  be  qualified  "L"  and
             nondetected target compound results should be qualified unusable (L).

       d.     If zero pesticide surrogate recovery is reported, the reviewer should examine
             the sample chromatogram  to determine if the surrogate may be present, but

-------
                                                        Procedure No.:  M2-PEST
                                                                     Revision:  1
                                                                Date: 06/30/1995
                                                                    Page 27 of 31


              slightly outside its retention time window.  If this is the case, in addition to
              assessing  surrogate   recovery  for  quantitative  bias,  the  overriding
              consideration is to investigate the qualitative validity of the analysis.  If the
              surrogate  is  not present, qualify  all nondetected  target  compounds as
              unusable (R).     .                             ••';••
                       *ป''•.•''••

 2.     If surrogate retention times in PEMs, individual standards, and samples are outside
       of the retention time limits,.qualification of the data is left up to the professional
       judgment of the reviewer.  Refer to Section 3.3.4.2 for more guidance.

 3.     Extreme or repeated analytical problems with surrogate recoveries should be noted
     •  for TPO action.   •; -••..=.•&##.••':' •;;-,• •'   .  • •       :  -.\V./  •'••:   '- .-:
 4.     Potential effects of the data resulting from surrogate recoveries not meeting the
       advisory limits should be noted in the data review narrative.
 3.8  Matrix Spikes/Matrix Spike Duplicates
'"."'•    •    '     '•";•••  :':'^^: ?•'•:•'' •'  .  ;      :
 Data for matrix spikes (MS) and matrix spike duplicates (MSD) are generated to determine
 long-term precision and accuracy of the analytical method on various matrices.  These data
 alone cannot be used to evaluate the precision and accuracy of individual samples.
 However, when exercising  professional judgment, MS/MSD data should be  used in
 conjunction with information on other deficiencies.

 3.8.1  Acceptance Criteria

 1.     Matrix spikes (MS) and  matrix spike duplicate (MSD) samples are analyzed at a
       frequency of at lest one MS and MSD per 20 samples of each matrix.

 2.     Matrix spike recoveries should be within the advisory limits provided on Form m
       PEST-1 and PESTV2 and in Attachment  1, Table 2.

 3.     Relative percent difference (RPD) between MS and MSD recoveries must be within
       the advisory limits provided on Form in PEST-1 and PEST-2 and in Attachment 1,
       Table 2.

 3.8.2  Review Items

 Form HI PEST-1 and PEST-2, and chromatograms

 3.8.3  Evaluation Procedures

 1.     Verify that MS and MSD samples were analyzed at the required frequency and that
       results are provided for each sample matrix.

-------
                                                        Procedure No.:  M2-PEST
                                                                      Revision:  1
                                                                Date: 06/30/1995
                                                                    Page 28 of 31
2.     Inspect results for the MS/MSD Recovery on Form ffl PEST-1 and PEST-2 and
       verify that the results for recovery and RPD are within the advisory limits.
4.     Check that the matrix spike recoveries and RPD were calculated correctly.

5.     Compare %RSD results of npnspiked compounds between the original result, MS,
       and MSD.

3.8.4 Action

1.     No action is taken on MS/MSD data alone. However, using informed professional
       judgment the data reviewer may use the MS- and MSD results in conjunction with
       other QC criteria and determine the need for some qualification of the data.

2.     The data reviewer should first try to determine to what extent the results of the
       MS/MSD affect the associated sample data.  This determination should be made.
       with regard to the MS/MSD sample itself, as well as specific analytes for all sample *
       associated with the MS/MSD.

3.     In those instances where it can be determined that the results of the MS/MSD affect
       only the sample spiked, then qualification should be limited to this- sample alone.
       However, it may be determined through the MS/MSD results that la laboratory is
       having a systematic problem in the analysis of one or more analytes,  which  affects
       all  associated samples.   For example, if  the recoveries for MS and  MSD are
       consistently low for both water  and soil samples, this could be indicative of a
       systematic problem in the laboratory  and  recoveries  should  be  examined  in all
       associated samples.

4.     The reviewer must use professional judgment to determine the need for qualification
       of positive results of nonspiked compounds.

       NOTE:      If a field blank was used for the MS/MSD, a statement to  that effect
                   must be included for the TPO.

3.9  Pesticide Cleanup Checks

Pesticide cleanup procedures are utilized to remove matrix interferences  from sample
extracts prior to analysis.  The use of the Florisil cartridge cleanup procedure significantly
reduces matrix interferences caused by polar compounds.  Gel permeation chromatography
(GPC) is used to remove high molecular weight contaminants that can interfere with the
analysis of target analytes.  Pesticide cleanup procedures are checked by spiking the

-------
                                                          Procedure No.: M2-PEST
                                                                       Revision:  1
                                                                  Date:  06/30/1995
                                                                     Page 29 of 31
cleanup columns and cartridges, and verifying  tne recovery of pesticides through  the
cleanup procedure.

3.9.1 Acceptance Criteria

1.     Florisil Cartridge Cleanup

       a.     Florisil cartridges must be used for the cleanup of all sample extracts.

       b.     Every lot number of Florisil cartridges used for  sample cleanup must be
             checked by  spiking  with  2,4,5-trichlorqphenol and  the  midpoint
             concentration of Individual Standard Mixture A.                '

       c.     The lot of Florisil cartridges is acceptable if the  recoveries  for all of the
             pesticides and surrogates in Individual Standard Mixture A are within 80 to
             120 percent, if the recovery of 2,4,5-trichlorophenol is less than 5 percent,
             and if no peaks interfering with the target analytes  are detected.

2.     Gel Permeation Chromatography (GPC)

       a.     GPC is used for the cleanup of all soil sample extracts and for water sample
             extracts that contain high  molecular weight components that interfere with
             the analysis of the target analytes.

       b.     At least once every 7 days, the calibration of the GPC units must be checked
             by spiking with two  check mixtures:  the matrix spiking solution and a
             mixture of 0.2 ug/ml Aroclors 1016 and  1260.

       c.     The GPC calibration is  acceptable  if the recovery of the pesticides in the
             matrix  spiking solution are within 80  to  110 percent,  and the Aroclor
             patterns should match those generated for previously run standards.

       d.     A GPC blank must be analyzed after each GPC calibration and is acceptable
             if the blank does not exceed one-half the CRQL for any target analytes.

3.9.2 Review Items

Form IX PEST-1 and 2 and chromatograms

3.9.3 Evaluation Procedure

1.     Florisil Cartridge Check

-------
                                                          Procedure No.: M2-EEST
                                                                       Revision: 1
                                                                  Date:  06/30/1995
                                                                     Page 30 of 31
       Check the data from the Florisil cartridge solution analyses and the Form DC PEST-
       1 and recalculate some of the percent recoveries to verify that the percent recoveries
      , of the pesticides and surrogates in Individual Standard Mixture A are within 80 to
       120 percent, the recovery of 2,4,5-trichlorophenol is less than 5 percent, and no
       interfering peaks are present.   Compare the raw data to the reported results and
       verify that no calculation or transcription errors have occurred.

2.     Gel Permeation Chromatography (GPC)

       Check the data from the GPC calibration check analyses and the Form DC PEST-2
       and recalculate some of the percent recoveries  to verify that the percent recoveries
       of the pesticides in the matrix spike solution are within 80 to 110 percent and that
       the Aroclor patterns are similar to those ofj>revious standards. Check to make sure
       that no transcription errors have occurred.

3.9.4 Action

1.     If Florisil Cartridge Check criteria are not met, the raw data should be examined'
       for the presence of polar interferences and professional judgment should be used to
       qualifying the  data.   If a laboratory  chooses to analyze samples under  an
       unacceptable Florisil Cartridge Check, then the TPO should be notified.

2.     If Gel Permeation Criteria are not met, the raw data should be examined for  the
       presence of high molecular weight contaminants and professional judgment should
       be used in qualifying the data. If a laboratory chooses to analyze samples under
       unacceptable Gel Permeation Criteria, then the  TPO should be notified.

3.     If zero recovery was obtained  for the pesticide compounds and surrogates during
       either check, then the nondetected target compounds may be suspect and  the data
       may be qualified unusable (R).

4.     If high recoveries (i.e, greater than 120, percent) were  obtained for the pesticides
       and  surrogates during  either check, use professional judgment to qualify  detected
       target compounds as biased high (K). Nondetected target compounds do not require
       qualification.

5.     Potential effects on the sample data result from the pesticide cleanup analyses  not
       yielding acceptable results should be noted in the data review narrative.

3.10  Reported  CRQLS

The objective is to ensure that contract required quantitation limits (CRQLs) are accurate.

-------
                                                        Procedure No. t  M2-PEST
                                                                     Revision:  1
                                                                Date: 06/30/1995
                                                                    Page 31 of 31
3.10.1 Acceptance Criteria
The adjustment of the CRQL, must be calculated according to the equations provided in
Attachment 1, Equations N and O.

3.10.2 Review Items

Form 1 PEST, Form X PEST-1 and PEST-2, and chromatograms.
                        k
3.10.3 Evaluation Procedure

1.     Verify  that  the  CRQLs  have been  adjusted  to reflect all  sample  dilution,
       concentrations,  splits,  cleanup activities and  dry weight factors  that are  not
       accounted for by the method. .
                      ' -          *" '      . '      "  .

3.10.4 Action

1.     If there  are  any discrepancies found, the laboratory may be contacted by  the
       designated representative to obtain additional information that could resolve any
       differences. If a discrepancy remains unresolved, the reviewer must  decide which
       value is the best value. Under these circumstances, the reviewer may determine if
       qualification  of the data is  warranted.   A description  of the  reasons  for data
       qualification and the qualification that is applied to the data should be documented
       in the data review narrative.

-------
                                                       Procedure No.:  M2-PEST
                                                                   Revision: 1
                                                              Date:  06/30/1995
                                                                  Page: 1 of 6
                         Attachment 1 to Appendix C-2
                   Multi-Media, Multi-Concentration
              Contractual Requirements and Equations
                        for Pesticide Data Review
                  GG/ECD Instrument Performance Check

Check the Performance Evaluation Mixture calculations using equations IL2, n.3, and n.4 to
ensure correct calculation of DT and Endrin breakdown. The breakdown of DDT and Endrin
in both of the PEM injections must be less than 20.0 percent, and the combined breakdown of
DDT and Endrin must be'.'less than 30.0 percent.               .   .

                              ^m found in ngfpDD^DDEtx 100        (A)
                                  'Amount in ng of DD7 injected
      % Breakdown Endrin =
                           . -        '                 •   •
            Amount found in  ng (Endrin aldehyde + Endrin ketone) x 100        ^\
                        Amount of Endrin injected in ng       '  . __
        Combined % Breakdown •- % Breakdown DDT + % Breakdown Endrin     (Q
All peaks in both the injections of the Performance Evaluation Mixture must be 100 percent
resolved on both columns. The relative percent difference of the calculated amount and the
true amount for each of the single component pesticides and surrogates in the PEMs must be
less than or equal to 25.0 percent using equation D.



                               =  F~"ป " Cซ*l x 100                       (D)
                                      nom
      Where:
                  True concentration of each analyte
                  Calculated concentration of each  analyte from the analysis of the
                  standard

-------
                                                            Procedure No.:  M2-PEST
                                                                         Revision:  1
                                                                   Date:  06/30/1995
                                                                        Page: 2 of .6
                                Initial Calibration
Retention time windows for each analyte and surrogate are calculated using Table 1. Windows
are centered around the mean absolute retention time for the analyte established during the
initial  calibration.   For example,  for a  given  pesticide the mean retention  time is first
determined from the initial calibration and found to be  12.69 minutes.   The retention time
window for this pesticide is ฑ0.05 minutes.  Therefore, the calculated retention time window
would  range from 12.64 to 12.74 minutes.               •  ..-',-.
Table 1
RETENTION TIME WINDOWS FOR PESTICIDE TARGET COMPOUNDS
Pesticide Compounds
alpha-BHC .;
beta-BHC
gamma-BHC
delta-BHC
Heptachlor .
Aldrin
aipha-Chlordane > ... '
gamma-Chlordane
Heptachlor epoxide
Dieldrin
Endrin
Endrin aldehyde
Endrin ketone
DDD
DDE
DDT
Endosulfan I
Endosulfan n
Endosulfan sulfate
Methoxychlor
Aroclors
Toxaphene
Tetrachloro-m-xylene
Decachlorobiphenyl
Retention Time Windows in Minutes
ฑ005
ฑ0.05
ฑ0.05
ฑ0.05
ฑ0:05
ฑ0.05
ฑ0.07
ฑ0.07
ฑ0.07 -
ฑ0.07
ฑ0.07
ฑ0.07
ฑ0.07
ฑ0.07
ฑ0.07
ฑ0.07
ฑ0.07
ฑ0.07
ฑ0.07
ฑ0.07
ฑ0.07
ฑ0.07
ฑ0.05
ฑ0.10
The %RSD of the calibration factors for each single component target compound must be less
than or equal to 20.0 percent.  The %RSD for the two surrogates must be less than or equal to
30.0 percent. Up to two single component target compounds per column may exceed the 20.0

-------
                                                           Procedure No.: M2-PEST
                                                                        Revision:  1
                                                                   Date: 06/30/1995
                                                                      'Page:  4 of.6
                             RPD
x 100
(I)
                                          nom
      Where:
             =    True concentration of each analyte
             =    Calculated  concentration  of each analyte from  the  analysis of  the
                   standard                  .          .      •   •     ~

                                Surrogate Spikes

The advisory limits for recovery of tetrachloro-m-xylene (TCX) and decachlorobiphenyl (DCB)
are 60 to 150 percent for both water and soil samples.  The surrogate percent recovery is
calculated using equation J.  The retention times of both surrogates must be  within  the
calculated retention time windows, i.e., TCX  must be within ฑ0.05 minutes of the  mean
retention time determined from the initial calibration and DCB must be within ฑ0.10 minutes
of the mean retention time determined from the initial calibration.     .                  .
                                                   O,
                       Surrogate Percent Recovery = --= x 100                    (J)
      Where:
             Qd = Quantity determined by analysis
             Q, = Quantity added to sample/blank
                     Matrix Spikes/Matrix Spike Duplicate

The matrix spike/matrix spike duplicate recovery and RPD requirements are listed in Table 2.
The matrix spike recoveries and RPD are calculated using equations K and L.
Table 2
MS/MSD CONTRACTUAL REQUIREMENTS
Compound
gamma-BHC (Lindane)
Heptachlor
Aldrin
Dieldrin
Endrin
4,4'-DDT
% Recovery
Water
56-123
40-131
40-120
52-126
56-121
38-127
RPD
Water
. 15
20 .
22
18
21
27
% Recovery
SoU
46-127
35-130
34-132
31-134
42-139
23-134
RPD
Soil
50
31
43.
38
45
50

-------
                                                          Procedure No.: M2-PEST
                                                                       Revision:  1
                                                                  Date:  06/30/1995
                                                                      Page:  3 of 6

percent limit for %RSD, but those compounds must have a %RSD of less  than or equal  to
30.0 percent.   Calibration factors are calculated using equations G and H and the %RSD  is
calculated using equations E and F.


                          %KSD = StandardDeviation f ^                      (E)
                                        Mean
      where:
                                            ฃ•*-ป!.••;.
                       StandardDeviation - |—	— f/2
      Where:.:.  '•" .-•..."•/''-.•••            .     '•  .."     f
             & - each individual value used to calculate the mean
             x = the mean of n values
             n = the total number of values
                     — _  Peak Area (V Height} of the Standard
                 ',                 Mass injected (ng)
                                         "   CF
                                  CF =     ^                               (H)
                                        i • i   n

      Where:
            CF    =     Mean calibration factor of n values
            CFj    =     i* calibration factor
            n     =     Total number of values
                            Continuing Calibration

The retention time (RT) for each target compound and surrogate must be within RT window as
calculated  above using  the mean absolute RT established  during  the three-point initial
calibration. The relative percent difference of the calculated amount and the true amount for
each of the compounds in the mid-point concentration of the Individual Standard Mixtures must
be less than or equal to 25.0 percent using equation I.

-------
                                                            Procedure No.:  M2-PEST
                                                                         Revision:  1
                                                                    Date: 06/30/1995
                                                                         Page:  5 of. 6
                           Spike Recovery = --— x 100                      (K)
       Where:
             SSR = Spike sample result
             SR = Sample result
             SA = Spike added   .
               ' >       ' .  •     ' .  •      •
                                    \MSR-MSDR\
                                  1/2 (MSR + MSDR)

       Where:'.-'     ''-''•;;;,: '•.'..  '•    '      :      '.  ...-. :
         -    RPD  =  Relative percent difference
             MSR  = Matrix spike recovery
             MSDR — Matrix spike duplicate recovery                .   ..

             The Vertical bars in the formula above indicate the absolute value of the
             difference, hence RPD is always expressed as a positive value.
                             Pesticide Cleanup Check

Every lot number of Florisil cartridges used for sample cleanup must be checked by spiking
with 2,4,5-trichlorophenol and the midpoint concentration of Individual Standard Mixture A.
The recoveries for all of the pesticides and surrogates in Individual Standard Mixture A must
be  within 80 to 120 percent, the recovery  off 2,4,5-trichlorophenol must be less than 5
percent, and no peaks must interfere with the target analytes. Percent recovery is determined
using equation M.

                                               Q.
                             Percent Recovery = —- x 100                        (M)
                                               ^ซ

      Where:
             Qd =  Quantity determined by analysis
             Q. =  Quantity added to sample/blank
The gel permeation chromatography (GPC) apparatus must be calibrated every 7 days.  The
calibration is acceptable if the recovery of each single component analyte is within 80 to 110
percent and the Aroclor patterns match patterns previously generated by standards.

-------
                                                       Procedure No.:  M2-PEST
                                                                    Revision:  1
                                                              Date: 06/30/1995
                                                                   Page:  6 of .6
                             ..Reported CRQLs
The CRQL of the single component pesticides is calculated using equations N and O, as
appropriate.                            .

CRQL for waters:

                      Sample CRQJL = Reference CRQLx Df                   (N)
                      •'• I •'-.:•
            Df    =     Dilution factor

CRQL for soils/sediments (dry weight basis):
                      Sample CRQL = Reference CRQL x                      (O)
                        100 - % moisture
                              100

-------
            Appendix D
Validation of Inorganic Analyte Data
              Level IM-1

-------
                                             Procedure No.:  IM-1-INORG
                                                            Revision:  2
                                                         Date: 06/30/95
                                                          Page:  1 of 26


    VALIDATION OF TARGET ANALYTE LIST METALS
                       AND CYANIDE DATA
                    MANUAL APPROACH BM-1

                 1.  PURPOSE AND APPLICABIIJTY

this procedure provides instructions  to manually validate the target analyte list (TAL)
metals and cyanide data using a manual innovative data validation approach that is based on
the EPA's National Functional Guidelines for  Data Review   and EPA Region Hi's
Modifications to the National Functional Guidelines for Data Review.  Specifically, the
approach is based  on the use of quality control (QQ information contained on. the
laboratory QC summary forms, and does not utilize' the raw data. The information that is
obtained from the QC summary.; forms is indicated ton Table D^WNORG-^.j^iThis
procedure is applicable to the TAL metal and cyanide data obtained using the (Hontract
Laboratory Program Statement of Work (CLP SOW).             -

Data validated using this procedure  are considered usable for the following types of
purposes; however, the data users must decide on a  case-by-case basis whether the
procedure is suitable for their intended data uses.  The suggested data uses are:

      •     Oversight of activities led by other parties

      •     Comparison to action levels

      •     Initial site investigation

      •     Contamination sources

      •     Nature and extent of contamination

      •     Preliminary risk assessment

      •     Risk assessment with known high levels of toxics

      •     Feasibility study

      •     Preliminary design

      ซ     Treatability study

      •     Initial cleanup verification


1M-1-INORO

-------
                                               Procedure No.: IM-1-INORG
                                                              Revision:  2
                                                           Date: 06/30/95
                                                            Page:  2 of 26

           2.  QUALITY CONTROL MEASURES CHECKED

Table IM-1-INORG-QC highlights the quality control (QC) indicators evaluated under this
data validation procedure.  These measures fall under one or more data validation steps
highlighted below.

2.1  PROCEDURES

Data validation is performed primarily with respect to the technical data quality criteria;
however, there are certain contractual criteria  that may reflect on  a participating
laboratory's compliance with the terms and conditions of the program and future audits of
the laboratory.  The term  "CCS"  next to a validation  step  indicates that  there is  a
contractual criteria in addition to the technical criteria.  It is incumbent upon a data
validator to point put the^cMtractUal deficiencies to-the laboratory's CLP Technical Project
Officer (TPO) for clarificMons and corrective action. The data validation steps, are as
follows:      '  ','-'->'fyฃ:V^X':.-i;:

      •     Action Level Notification

      •     Technical Holding Times (CCS - Contractual holding times only)

      •     Calibration

            •     Initial (CCS)

            •     Initial and Continuing Calibration Verification (CCS)

            •     CRDL Standards

      •     Blanks

            •     Initial Calibration Blank (CCS)

            •     Continuing Calibration Blank (CCS)

            •     Preparation Blank (CCS)

            •     ICP Interference Check Sample (CCS)

            •     Laboratory Control Sample (CCS)

            •     Duplicate Sample (CCS)
1M-MNORO

-------
                                                Procedure No.:  IM-1-INORG
                                                                Revision: 2
                                                             Date:  06/30/95
                                                              Page: 3 of 26
Table IM-1-INORG-QC
QC CHECKLIST FOR LEVEL IM-1
CLP RAS TARGET ANALYTE LIST INORGANICS
1 • • ' ' . . • • ' • • 1 .. • • ' • • '
'..< QC MEASURES
' . •' • •• • . . • '•••'• 1
Action Level Notification .
Holding Times
v
Initial Calibration
Initial Calibration Verification / . ^
Continuing Calibration Verification • • ' • •
..**.. ' v • • •. ' . . ' . ' _ •


Laboratory Blank '.•

Kjf interterence mecK • . .
Laboratory Control Sample
Duplicate Precision
Matrix Spike Recovery •
Furnace QC (MSA)
ICP Serial Dilution
Field Duplicates
Reporting Limit Verification
Sample Paperwork ,
Raw Data
| IM-1 |
|: . x \
x
X
.x/ '"•'• .
• ' • • x' : " ". '•
•••-.-.-'•xrj
• . ';.X; ::.; '
x
X
X
X
X
X
X
X
X
^M
M2
1 x
x
1
X
X
•' "' -.-x .. . •
,-:•• ." x :
X
x
^^ X
X
X
X
X
X
X
X
X
1M-1-D40RG

-------
                                                Procedure No.:  IM-1-INORG
                                                                Revision:  2
                                                             Date:  06/30/95
                                                              Page: 4 of 26

       •      Matrix Spike Sample^CCS)

       •      Graphite Furnace Atomic Absorption QC (CCS)
                        v               •
       •      Method of Standard Addition (MSA)                       .

       •      ICP Serial Dilution (CCS)
               . .   '! .   .     '•'..••'••.     '"..•'•      '        •;
       •      Field Duplicates (if included)
             "•  .  \ '          '''''•'.   ' ••'       '  '       ป.•",•.'

       •      Reporting Limit Verification (IDLs,  linear range, dilution factors and
             moisture content)
           I • •     • :      .-,'-•     .        .        '             -      •
       •      Sample Paperwork             ^-?:

     REPORTING REQUIREMENTS

Several forms have been developed to assist in the performance evaluation and in keeping;
track of the data quality qualifiers:  The first form, IM-1-INORG-HT,  summarizes the'
holding times.   The  second form, IM-1-INORG-CAL allows  documentation of the
calibration and blank QC violations.  The third form, IM-1-INORG-SPK is used for the
spike recoveries, duplicate precision, and control sample analyses data.  The fourth form,
IM-1-INORG-QL is used for summarizing all qualifiers for the samples. The EPA Region
HI Inorganic Regional Data Assessment form is used by  the data validator to summarize
contractual deficiencies for the laboratory's CLP Technical Project Officer (TPO).   All
forms are appended at the back of this SOP.

A memorandum describing those elements that were outside of established  QC criteria, the
actions which were taken and the impact on  data usability  must be prepared  with
substantiating documentation.  The report-and the supporting documentation should include
the following:

       1.     Hand annotated Form Is with
             a.     data validation qualifiers         •
             b.     sample identification number
             c.     sampling location

       2.     A narrative description with
             a.     a statement that defines the level of the data review,  i.e., IM-1
             b.     major and minor problems associated with the analysis
             c.     issues that may have affected detection limits
1M-1-INORO

-------
                                                Procedure No.:  IM-1-INORG
                                                                Revision:  2
                                                             Date:  06/30/95
                                                              Page: 5 of 26

       3.     The following attachments
             a.     list of data validation qualifiers
             b.     documentation which includes forms that support the  assigned data
                   qualifiers.   Samples  affected;  by  noncompliant  quality control
                 . measures should be listed on these forms.
             c.     chain of custody form(s)
GLOSSARY OF DAT A QUALIFIER CODES (INORGANIC)

The codes  described below are (hose recommended in the EPA's national Functional
Guidelines and the Region ffl's modifications.
Codes Relating to •
(confidence concerning presence or absence of analytes):
U       •=' ':'.'•'  ':•   r >^ :iileปciid. ^The associated number  indicates approximate
                      sample'ooncentration necessary to be detected.'
           . .  •  .    ••'• .'"• _ >:•••"• v-fe^v-^fcv'-v-Vv.-, "••':: ••-'   :..'.'••     •      :.•:•••'•

(NO CODE) =        Confirmed identification.

B        =           Not detected substantially above the level reported in laboratory
                      or field blanks.                            - -Z^

R        =           Unreliable result.  Analyte may or may not be present in the
                      sample.  Supporting data necessary to confirm result.

Codes Relating to Quantitation
(can be used for both positive results and sample quantitation limits):

J         =           Analyte present. Reported value may not be accurate or precise.

K        =           Analyte present.  Reported value may be biased high.  Actual
                      value is expected lower.

L        =           Analyte present.  Reported value may be biased low.  Actual
                      value is expected to be higher.

UJ       =           Not detected, quantitation limit may be inaccurate or imprecise.

UL       =           Not detected, quantitation limit is probably higher.
IM-l-INORO

-------
                                                  Procedure No.:  IM-1-INORG
                                                                  Revision: 2
                                                              Date:  06/30/95
                                                                Page: 6 of 26

 Other Codes

 Q        =        ;   No analytical result.

 *         =           Results reported from diluted analysis.
 The following subsections describe for each of the QC indicators die acceptance criteria,
 location and retrieval of QC data, evaluation of the QC data, actions taken in the event the
 QC acceptance  criteria are exceeded, and documentation of the QC  violations  in a
                                    '                     ' '
 2 3  ACTION LEVEL NOTIFICATION
  '-..••'."''  •'  '  •   .ii,",- .' '.'•  •'.'!.•' '.'-.. "••.'•'' '••".    • ,.  ' •. •  •  ... •'.<••  . .' *•.ซ=ป--••..• :: ..-'.. .'.-. '.:.- • • •..'
 The purpose behind action level notification is to make the EPA Remedial Project Officer
 (RPM) or the Site Project Officer (SPO) aware ef the potential human health risk at the
 site.  In accordance with me Region "ID Hazardous Waste Division policy, the EPA RPM
 or SPO must be promptly notified of any contaminant exceeding the established action level
^or the 10-day health advisory limit The data for contaminants exceeding the action levels
 must be validated as a top priority and reported to the RPM or SPO, as soon as possible. •
 Validation of the rest of the data may then be completed within the normal time frame.

 2.3.1  Acceptance  Criteria
     *  i        ..''.'-."-•'                                 '  •  *    .
 EPA's Office of Solid Waste and Emergency Response has established -10-day advisory
 limits and action levels for several organic compounds and elements of special health risk
 concerns based on the Safe Drinking Water Act  The target analytes and their 10-day
 health advisory limits are listed in Table IM-1-INORG-AL.  The  criteria for action level
 notification are as follows:

       •      The contaminant concentration must be equal to or above the established 10-
              day health advisory limits.

       •      Data for contaminants exceeding the action levels must be validated as a top
              priority.

       •      The following  EPA personnel  must be notified  of the  action  level
              exceedances:

              •      EPA RPM or SPO
              •      EPA Section Chiefs

       •      The remaining data validation should be completed per normal procedures.
 1M-1-INORO

-------
                                                 Procedure No.: 1M-1-INORG
                                                                  Revision:  2
                                                               Date:  06/30/95
                                                                Page: 7 of 24

             Any special instructions from the  Hazardous Waste  Division should  be
             followed.

             Records should be kept of the data review, action level notification and any
             follow up instructions and actions.
Table IM-1-INORG-AL
TARGET ANALYTE LIST METALS AND ACTION LEVELS
Compound "
Arsenic, •-• '• -•; W- ... .:••••
.. .1 .':.'.'
Chromium - :
Nickel
'Action
Level'
-.x;'--,5d-v- .'••
1000
- .^1000 •'-:•<;
Compound .
Cadmium -' •

Lead'" '" ''• '' '•': ,'''• -
Cyanide ,,.. :••••'-• ''••'••'-.
' ' Action
Level'
yr>/r'.40.
v -.-?•• if'- ••
200
' All units are ug/1. '•• ' . . . :
- ... - .- ^. • r .'"•'-' -J • - ' ' ". '.'.'' ' ' ' - * '
b Lead also has an action level of 500 mg/kg (ppmw) for soil samples.
2.3.2 Data requirements and Retrieval of Data               ,_
                 ' '            •               *                   * *^-*.
All forms required to perform Level IM-1 validation, as detailed in the following sections
are necessary for carrying out action level notification.

2.3.3 Evaluation Procedure

The evaluation process preceding action level notification will primarily  consist of
comparing the results on Form Is with the action levels presented in Table IM-1-INORG-
AL. Following the identification of the contaminants exceeding the action levels, focused
data validation should be performed using the criteria,  and procedures described  in the
appropriate sections below.

2.3.4 Action

The action resulting from focused data validation will be the notification of action level
exceedance to the personnel identified above in Section 2.3.1. Copies of Form Is can be
used to highlight the contaminants above the action levels.   The findings of the focused
validation can be  summarized in a memorandum, and  the data  qualifiers resulting from
focused validation may be written on the Form Is. The marked up forms should state that
they represent validation of only the contaminants exceeding the  action levels, and not all
data.
1M-MNORO

-------
                                                Procedure No.:  1M-1-INORG
                                                                Revision:  2
                                                             Date: 06/30/95
                                                              Page:  8 of 24
2.4 HOLDING TIMES
The primary objective is to ascertain the validity of results based on the holding time of the
sample from time of collection to time of sample extraction and analysis.  The secondary
objective is also to verify  compliance with the contractual extraction and analysis holding
times from the verified time of sample receipt (VTSR> at the laboratory.
2.4.1  Acceptance Criteria;

Technical  requirements for sample holding times have only been established for water
matrices.   The  following holding times  (from the time of  sample collection)  and
preservation requirements were established under 40 CFR 136 (dean Water Act), and are
found:in  Volume49, Number209 of  the  Federal Register,  page43260,  issued on
October 26V 1984i  \ ;'•'"      ~
>.-^. -,^.:I<*L<; —,.
    •'•."/•   :  Metals:        6 months; preserved to pH < 2
       •     Mercury:     28rdays; preserved to pH < 2
       •     Cyanide:      14 days; preserved to pH > 12

Contractual holding times have been established by the CLP for the water and soil/sediment
samples.  The times are counted from the time of sample receipt at the laboratory, and are
as follows:

       •     Metals:      180 days
       •     Mercury:     26 days
    •   •     Cyanide:      12 days

Generally, the holding times are calculated using the dates only and not the times.

2.4.2 Data Requirements and Retrieval of Data

Forms  1, 13, 14, and EPA Traffic Reports and Sample Shipping Logs.

2.4.3  Evaluation Procedure

Technical holding times  are established  by comparing the sampling  date on the EPA
Sample Traffic Report  with the dates of sample preparation/extraction and analysis found
on the extraction and instrument run logs (Forms  13 and 14, respectively).

           Technical Holding Time (days) = Analysis Date - Sampling Date

Contractual holding times are established by comparing the sample receipt date on Form Is
with the sample preparation/extraction and analysis dates on the extraction and.run logs
(Forms 13 and 14, respectively).
1M-1-INORO

-------
                                                  Procedure No.: 1M-1-INORG
                                                                   Revision:  2
                                                               Date:  06/30/95
                                                                Page: 9 of 24

           Contractual Holding Time (days) = Analysis Date - Receipt Date

2.4.4 Action

The following actions are required only for the violations of the technical holding times. If
contractual holding times are exceeded, then the TPO must be notified of the violation.

1.     If 40 CFR 136 criteria for technical holding times and preservation are not met,
       qualify all results > Instrument Detection Limit (DDL) as biased low (L) and the
       results < IDL as estimated detection limits biased low (UL).

2.     If holding times are exceeded  by two times  (2X) the criteria  for mercury  and
       cyanide, qualify the non-detected results (results < IDL) as unusable (R).

3. •    If there are gross violations of the holding times for the metals, use professional
       judgment to determine the reliability of the data. A low bias would be expected for
       significantly longer holding times, and the reviewer may reject  the non-detected
       data (results < IDL) as unusable (R).
                          ..  v  ''•••'       '            ••"'..'
4.     Although the  technical holding  times for soil samples have not  been established,
       apply the water holding, time criteria to the soil samples. If the soil sample data are
       qualified using the water holding time criteria,  this must be clearly "documented in
       the narrative report.

2.5  CALIBRATION

Requirements for satisfactory instrument calibration  are established to ensure that the
instrument  is capable  of producing acceptable quantitative  data.   Initial  calibration
demonstrates  that the instrument is capable of acceptable performance at the beginning of
the analysis  run,  and  continuing  calibration   verification  documents  that  the initial
calibration is  still valid.

Standards at concentrations near the lower limit of detection are also required to be run to
determine the linearity of the instrument.

2.5.1  Acceptance Criteria

1.     Initial  Calibration.    Instruments must  be calibrated  daily  and each time  the
       instrument is set up.  Specific requirements for each type of analysis are as follows:

       •      ICP Analysis

             •      A blank and at least one standard must be used in establishing the
                    analytical curve.

-------
                                                 Procedure No.: 1M-1-INORG
                                                                  Revision:  2
                                                               Date:  06/30/95
                                                               Page:  10 of 24
             Atomic Absorption Analysis (AA)
             •     A blank and at least three standards, one of which must 'be at the
                   Contract Required  Detection  limit (CRDL),  must  be used  in
                   establishing the analytical oirve.

       •      Mercury Analysis ';".: ',*.-•'.'         '

             •     A blank and at least four standards must be used in establishing the
                   analytical curve.   -.',  -'
             •     The correlation coefficient must be  ^ 0.995.  (This is a technical
            •    .'critmon and not a contractual one;)

       •      Cyanide Analysis

       _,  ,   •  .   A blank and at least three standards must be used in establishing the
                   analytical curve.     ;                            •           ••   ซ
             •     The correlation coefficient must be  ^ 0.995.  (This is a technical
                   criterion and not a contractual one.)

2.     Initial and Continuing Calibration Verification flCV and CC\0.

       •      Analysis  results must fall within the  control limits  of 90 to 110 percent
             Recovery  (%R) of  the  true value for all analytes  except  mercury and
             cyanide.

       •      Analysis  results for mercury must fall within  the control limits  of 80 to
             120%R.

       •      Analysis  results for cyanide  must fall within the control limits  of 85  to
3.    CRDL Standards for ICP (CRT) and AA (CRAV

      •      A  CRI must be run at  a concentration  of 2X CRDL, or 2X the IDL,
             whichever is greater, for each ICP analyte (except Al,.Ba, Ca, Fe, Mg, Na
             and K) at the beginning and end of each sample run or a minimum of twice
             per 8 hours.

      •      A  CRA must be run at a concentration equal to the CRDL,  or the IDL,
             whichever is greater, at the beginning of each sample run.
1M-1-INORO

-------
                                                 Procedure No.: IM-1-INORG
                                                                Revision:  2
                                                             Date: 06/30/95
                                                             Page:  11 of 26

       •     The CRDL standard recoveries should be between 90-110% of the "true
             values.,

2.5.2 Data Requirements and Retrieval of Data

Forms 1, 2A, 2B arid 14.          .

2.5.3 Evaluation Procedure

1.     Verify that the instrument was calibrated daily and each time the instrument was set
       up using the correct number of standards and blank.            ^?>"i-      ;
2.    Recalculate one or more of the ICV and CCV  %R per type of analysis (ICP,
      GFAA, etc.) using the following equation and verify that the recalculated value
      agrees with the laboratory reported values on Form 2A.  Due to possible rounding
      discrepancies, allow results to fell within 1 percent of the contract Windows (e.g.,
      80 to 111 percent),

                              %R = Found x 1QO
                                     True

where:
      Found = Concentration (in  ug/L) of each analyte measured in the analysis of
             the ICV or CCV solution                       v

      True   = Concentration (in ug/L) of each analyte in the ICV or CCV source

3.    Verify from the run log (Form 14) that the CRAs  and CRIs were run at the
      required frequency.

4.    Verify from Form 2B that the CRIs were at 2X CRDL, or 2X IDL, whichever, was
      greater, and the CRAs were at the CRDL, or the IDL, whichever was greater.

2.5.4  Action

1.    If the minimum number of standards as defined  above  were not used for initial
      calibration, use professional judgment in qualifying  the data.  However, if the
      instrument was not calibrated daily and each time it was set up, qualify the data as
      unusable (R). Document the noncompliance with the calibration requirements in the
      narrative and document on the EPA Region III Inorganic Regional Data Assessment
      Form.
1M-1-INORQ

-------
2.
                                                Procedure No.: IM-1-INORG
                                                                Revision: 2
                                                             Date: 06/30/95
                                                             Page:  12 of 26

      If the ICV or CCV %R falls; outside the acceptance windowns, using the following
      guidelines for qualifying the data.
Percent Recovery (%R)
Metals
90-110
75-8?
>110
<75
Cyanide
85-115
,70-84 '
>115
<70 ::
Mercury
J; 80420
:'^;'65V79';' .
'<ฃ >120 ; ;
3-;<ฅ-:. :
Qualifier
>IDL
None
Vv--L-V:'ซ:>
.•-:.-•". K ......
'.-•' L. •/.••/•
 IDL as biased low (L), and results < IDL as biased low (UL).
      >      If the ICVor GGV %R is >110 (metals) or     5;20
          :   results < IDL are acceptable.  Results > IDL should be qualified as biased
             high(K).    >

      •      If the IGV or CCV %R is <75 percent (medals) or <70j)ercent (CN) or
             < 65 percent (Hg), qualify all positive results as unusable (R).

      Please be advised, there are no National Functional Guidelines to qualify the data
      based on the CRDL analyses; however, EPA Region HI has developed specific
      guidance to qualify the data.

      •      If the recovery for the CRI or CRA is >110% and the reported sample
             results are >EDL, but < 2X CRDL, qualify the data as biased high (K).

      •      If the recovery for the CRI or CRA is between 50-89%, qualify results >
             IDL, but <  2X CRDL as biased low.  Qualify results < IDL as biased low
             (UL).

      •      If the recovery for an element is <50%, qualify the results  > IDL, but <
             2X CRDL as biased extremely low (L). Qualify results < IDL as unusable
             (R).
1M-1-INORO

-------
                                                   Procedure No.:  IM-1-INORG
                                                                    Revision:  2
                                                                Date:  06/30/95
                                                                Page:  13 of 26
 2.6  BLANKS
 Blank results are assessed to determine the existence and  magnitude of contamination
 problems.  The criteria for evaluation of blanks applies to any blank associated with the
 samples.  If problems with any, blank exist, all data associated with the Case must be
 carefully evaluated to determine whether or not there is an inherent
 variability in the data for the Case, or if the problem is an isolated occurrence not affecting
 • .• •   fl *    V .•  -    -..-...-'  .• , ;.'.;&&,  .'-,"..   (• ,                .-    '   -       ' • •     •  •
 other data.   .                  •

 2.6.1  Acceptance Criteria

 No contaminants should be in the blank(s) at concentrations, > IDL.

 2.6'. 2  Data Requirements and Retrieval of Data
.,'•"•-  :- ' -•  .- .  -' •"'•-. - '.', ,'•-;'••* • '< \*-*?ri* •"1*:"^*'' V't; ' -     • •*       •  -
 •: •••'•• :>-\--ซ--.^>-V1:-:y-':v:'.->^i:-^-v.- •-• v'  •: .
 Forms land 3.          ;   .;   ;

 2.6.3  Evaluation Procedure
         • '  -        • .  •     \       '   •       '       -       .  •
                           ,,      .   -           _         ,   •.  . .
 Review the results reported on the Blank Summary (Form 3) for all blanks.  Note that the
 instrument blanks serve two purposes:  One, to determine any sample -carryover, and
 second, to determine the shift in the instrument baseline.  It is common that the instrument
 baseline may shift upward or  downward  from the calibration reference point  (zero
 concentration standard) during a sample run. This shift is not necessarily indicative of any
 carryover from the previous sample.  Negative blank results are common indicating a shift
 in the baseline.  Professional judgment should be used when qualifying data based on the
 instrument blanks.

 2.6.4  Action             v

 Action in the case of unsuitable blank results depends on the circumstances and origin of
 the blank.

       •     Sample results > IDL but < five times the amount in any blank should be
             qualified as (B).

       •    ' Any blank with a negative result whose absolute value is > CRDL must be
             carefully evaluated to determine its effect on the sample data.

       •     Qualify the field blanks in a  manner similar to that for samples.  A field
             blank may not be used to qualify another field blank.
1M-MNORO

-------
                                                 Procedure No.: IM-1-INORG
                                                                Revision:  2
                                                             Date: 06/30/95
                                                             Page:  14 of 26
                                               <           i
             When multiple field blanks are present in a SDG, use the highest detected
             concentration  for each analyte in the field and laboratory blanks when
             qualifying the sample data.

Note:  The blank analyses may not involve the same weights, volumes, or dilution factors
as the associated samples.  In particular, soil sample results reported oh Form 1 will not be
on the same basis (units, dilution) as the calibration blank data reported on Form 3.

2.7 ICP INTERFERENCE CHECK SAMPLE          '    ; ;t t J

The ICP Interference Check Sample (ICS) verifies the contract laboratory's interelement
and background correction factors.                                    ^

2.7.1  Acceptance Criteria                •*-'.                    ^;  ' ^y

1.     An ICS must be run at the beginning and end of each sample analysis run (or a
       minimum of twice per 8-hour working shift, whichever is more frequent).
•"' •'   •   •    •; ".   '.  • ' .../•,'  '.'  .  .   .'   '.      '.       . '•"  .='•  ••.V*1*- .;•'•;  :  '-  .
2.     Results for the ICS solution AB analysis must fall within the control limits of ฑ "
       20 percent of the true value. ,

2.7.2  Data Requirements and Retrieval of Data

Forms 1 and 4                   '•-.'"

2.7.3  Evaluation Procedure~  ....

1.     Verify at random the reported %Rs for the  Solution AB using the true and found
       values.  Use the following equation:

                    ,  ,GS %R m  Found-Solution AB  x  m
                                 True. Solution AB

where:
     Found Solution AB   =  Concentration (in ngfl) of each analyte measured in the
                            analysis of solution AB    v

      True'Solution AB   =  Concentration (in /ug/1) of each analyte in solution AB
                    i1    .            '

2.     Check the results with an absolute value  >  DDL for those analytes which are not
       present in the ICS solution.
1M-1-INORO.

-------
                                                   Procedure No.:  IM-1-INORG
                                                                    Revision: 2
                                                                 Date:  06/30/95
                                                                 Page:  15 of 26
2.7.4 Action
1.     For samples with concentrations of aluminum, calcium, iron, and magnesium which
       are comparable to or >  their respective levels in the Interference Check Sample:

       •     If the ICS recovery for an element is >  120 percent and the sample results
             are < IDL, this data is acceptable for use.                     .

       •     If the ICS recovery for an element is >  120 percent and the sample results
             are > IDL, qualify the affected data as biased high (K).

       •     If the ICS recovery for an element falls between 50 and 79 percent and the
           -  sample results are >  IDL, qualify the affected data as biased low (L).
         '.  •   .'"•'... •••''''-.•  -v5iN^''' •••"•'"' ••'•'"• ''-':'  "•"*.••.'•       •   .'••'•  .'"•'•    .   ''     '  '
       •If sample; results are <  IDL,  and  the ICS recovery for  mat analyte  falls
             within the range of 50 to 79 percent, the possibility of false negatives  may
             exist.  Qualify the data for these samples as detection limits biased  low
       •     If ICS recovery results for an element is <  50 percent, qualify results >
             IDL as biased low (L)r and results <  DDL as unusable (R).

2.   ซIf results  >  IDL are observed for elements  which. are not present in the ICS
       solution,  the possibility of false positives exists.  An evaluation of the associated
       sample data  for  the affected  elements  should be made.    For samples with
       comparable or higher, levels  of interferents.. and., with. analyte  concentrations that
       approximate those levels found in the ICS (false positives), qualify sample results
       > IDL as biased high (K).

3.     If negative results are observed for  elements that are not present in the  EPA ICS
       solutions, and their absolute value is  > 'IDL, the possibility of false negatives in the
       samples may exist.  If the absolute value of the negative results is > IDL, an
       evaluation of the associated  sample data should be made.   For  samples with
       comparable or higher levels of interferents, qualify results for the affected analytes
       <  IDL as biased low (UL), qualify results for the affected analytes > DDL as
       biased low (L).

4.     In general, the  sample data can be accepted if the concentrations of aluminum,
       calcium, iron, and magnesium in the sample are found to be < or equal to their
       respective concentrations in the ICS.  If these elements are present at concentrations
       > the' level in the ICS, or other elements are present at concentrations > the level
       in the ICS, or other elements are present in the sample at > 10 mg/L, the reviewer
       should investigate the possibility of other interference effects by using  the Table
       found in  the most recent  version  of the SOW.   These analyte  concentration

1M-MNORO

-------
                                                Procedure No.:  IM-1-INORG
                                                                Revision:  2
                                                             Date:  06/30/95
                                                             Page:  16 of 26

       equivalents presented in the: table should be considered only as estimated values,
       since the exact value of any analytical system is instrument specific.  Therefore,
       estimate the concentration produced by an interfering element.  If the estimate is
       >2X CRDL and also  > 10 percent of the reported concentration of the affected
       element, qualify the affected results as biased high (K).

2.8 LABORATORY CONTROL SAMPLE

The laboratory control sample (LCS) serves as a monitor of the overall performance of all
steps in the analysis, including the sample preparation.

2.8.1  Acceptance Criteria

1.     All aqueous LCS results, must fall within the •control limits of 80 to 120%R, except
       Sb and Ag which have no control limits.
2.    All solid LCS results must fall within the control limits established by the EPA.
      This information is available from EMSL/LV.

2.8.2  Data Requirements and Retrieval of Data

Forms 1 and 7.

2.8.3  Evaluation Procedure

1.    Review Form- VII and verify that results fall within the control limits.

2.    Randomly verify the reported recoveries on Form VQ using the following equation:
                         LCS %R -      Found  x 100
                                     LCS True

where:
   LCS Found      .=     Concentration (in /zg/1 for aqueous; mg/kg for solid) of each
                         analyte measured in the analysis of LCS solution

    LCS True .     =     Concentration (in /tg/1 for aqueous; mg/kg for solid) of each
                         analyte in the LCS source

2.8.4  Action

1.    Aqueous LCS:
1M-1-INORO

-------
                                                 Procedure No.: IM-1-INORG
                                                                 Revision: 2
                                                              Date:  06/30/95
                                                              Page:  17 of 26

             If the LCS recovery for any analyte falls  within the range  of SO  to
             79 percent, qualify results > IDL as biased low (L). If the LCS recovery is
             > 120 percent, qualify results > IDL as biased high (K).

             If results are < IDL and the LCS recovery is > 120 percent, the data are
             acceptable.
             If results are < IDL and LCS recovery  Ms within the range of SO to
             79 percent, qualify the data for the affected analytes as biased low (UL).

             If LCS recovery results are < . 50 percent,  qualify results > DDL as biased
             low (L), and results  < IDL as unusable (R).
2.    Solid LCS:                                                     .

      •     If the solid LCSlre^oyery^ibr any analyte is below the lower acceptable
             control limit, qualify all sample results > IDL as biased low (L).  If the
             LCS recovery is above the upper control limit, qualify the results > IDL as -
        '.    biased                        '   "''         ''                "
       •      If the LCS results are lower than the control limits, qualify all sample
             results  < IDL as detection limits biased low (UL).   '

       •      If the LCS results are higher than the control limits and the^ sample results
             are < IDL, the data are acceptable.

2.9  DUPLICATE SAMPLE ANALYSIS

Duplicate analyses are used to determine the laboratory precision for each sample matrix.

2.9.1  Acceptance Criteria

1.     Samples identified as field blanks cannot be used for duplicate sample analysis.

2.     A control limit of ฑ20 percent (35 percent for soil) for the Relative Percent
       Difference (RPD) shall be used for sample values >5X CRDL.

3.     A control limit of ฑ CRDL (ฑ2X CRDL for soil) shall be used for sample values
       < five times  CRDL, including the case when only one of the duplicate sample
       values is <  five times CRDL.
1M-1-INORG

-------
                                                 Procedure No.:  IM-1-INORG
                                                                 Revision: 2
                                                             Date:  06/30/95
                                                             Page:  18 of 26
2.9.2 Data Requirements and Retrieval of Data

Forms 1 and 6.

2.9.3 Evaluation Procedure

1.  ...  Review Form 6 and verily that results fall within the control limits.
    ' '                                                            '
2.     Recalculate one or more RPD using the following equation to verify that results
       have been correctly reported on Form 6.
where: "     ,.-••-.    -    • -               .
     S  '•••'—  First sample value (original)            >
     D    =  Second sample value (duplicate)

3.    Verify that the field blank was not used for duplicate analysis.

2.9.4 Action

1.    If duplicate analysis  results for a particular analyte fall outside the appropriate
      control windows, qualify results > IDL, for that analyte in all associated samples
      of the same matrix as estimated (J), and results < IDL as estimated (UJ).
             i
2.    If the  field blank  was used for duplicate analysis, all other QC data must be
      carefully checked and professional judgement exercised when  evaluating the data.
      Document this information on the  EPA Region  m Inorganic  Regional Data
      Assessment Form.                  -

2.10 MATRIX SPIKE SAMPLE ANALYSIS

The  matrix spike sample  analysis provides information about the effect of each sample
matrix on the digestion and measurement methodology.

2.10.1 Acceptance Criteria

1.    Samples identified as field blanks cannot be used for spiked sample analysis.
1M-1-INORO

-------
                                                  Procedure No.:  IM-1-INORG
                                                                  Revision:  2
                                                              Date:  06/30/95
                                                              Page:  19 of 26

 2.     Spike recovery (%R) must be within the limits of 75 to 125 percent  However,
       spike recovery limits do not apply when sample concentration  exceeds the spike
       concentration by a factor of four or more.

 3.     If the matrix spike recovery limits are not met and the. sample concentration does
       not exceed 4X the spike added, a post-digestion spike must be  performed for the
       ICP, CN and flame AA analyses.  This criteria does not apply 'to silver. The post
       digestion spike must be performed at 2X the sample concentration or 2X CRDL,
       whichever is greater. •:ฃ <%:

 2.10.2 Data Requirements and Retrieval of Data
2 J0.3 Evaluation Procedure

1.   '  Review .Form 5A and verify that results fall within the specified limitsV _;;.-.

2.     Recalculate one or more %R using the following equation to verify that results were
       correctly reported on Form 5A.


                             %R =  125 percent and the reported sample results are < IDL,
       the data is acceptable for use.

2.    If the spike recovery is > 125 percent and the sample results are >  IDL, qualify
      the data for these samples as biased high (K).
1M-1-INORO

-------
                                                 Procedure No.:  IM-WNORG
                                                                 Revision:  2
                                                              Date:  06/30/95
                                                              Page:  20 of 26

3.     If the spike recovery is < 75; percent and the sample results are > IDL, qualify the
       data for these samples as biased low (L).

4.     If the spike recovery .falls within the range of 30 to 74 percent and the sample
       results are < IDL, qualify the data for these samples as detection limits biased low
                                            '''      '    '
5,    If spike recovery results fall <  30 percent and the sample result .are < IDL,
      qualify  the data for these samples .as unusable (R).  If the results are > IDL,
      qualify the data as biased extremely low (L).

6.    If the field blank was used for matrix spike analysis, all other QC data must be
      carefully checked and professional judgement exercised when evaluating the data.
      Report this information on the EPA Region HI Inorganic Regional Data Assessment
      • form.     .•            ,       '.'.  '          .          ...  -'..:.-; i; •. .•' .. .

7.    If a post-digestion spike ^was not performed when required, or the recoveries fall
      outside me matrix spike recover
      digestion recovery data are not used for data qualification as acceptable recovery •
      limits have not been established by the EPA yet.)
                   •       .-. \  "  -, • "     '       (                   • -    '„••*••

2.11 FURNACE ATOMIC ABSORPTION QC

Duplicate injections and post-digestion/analytical spikes are required for each element and
sample analyzed by the furnace atomic absorption technique. Additionally, a reanalysis by
the method of standard addition (MSA) is required for samples meeting certain conditions
specified in the most recent version of the SOW.   The  post-digestion/analytical spike
recoveries for the GFAA metals are reported on Form 14 under the  %R column.  The
correlation coefficient (r) can be found on Form 8 for samples requiring MSA. The  flag
"+"  is placed on Form Is for samples and analytes with r <0.995.

2.11.1  Acceptance Criteria

1.    The post-digestion/analytical spike recovery for the GFAA metals must be within
      85-115%..

3.    The correlation coefficient (r) for the samples requiring reanalysis by the MSA must
      be 0.995 or better using the ordinary least squares linear regression. (A formula for
      linear regression can be found in the SOW, or some computer software applications
      have the linear regression function built-in.)
1M-MNORO

-------
                                                 Procedure No.:  IM-1-INORG
                                                                 Revision: 2
                                                             Date:  06/30/95
                                                             Page:  21 of 26

2.11.2 Data Requirements and Retrieval of Data

Forms 1, 8 and 14.

2.11.3 Evaluation Procedure

1.     Evaluate the %Rs from  Form 14 for a comparison with the analytical  spike
   .    recovery requirement of 85-115%.
                      ..'••.''.''•         ~ .
2.     Verify by random recalculation that the r values reported on Form 8 are accurate
       and equal to or greater than 0.995.

2.11.4 Action

1.     If the analytical spike recovery is less than 85%, but greater  than 40%, qualify
       results > IDL as biased low, (L), and results < IDL as biased  low (UL).    "

2.     If the analytical spike recovery is greater than  115%, qualify  results  >  IDL as
       biased high (K),  sample results  < IDL are acceptable.

3.     If the analytical spike recovery is less than 10% and the results are < IDL, the data
       should be qualified usable (R).  Sample results > IDL should be qualified biased
       extrememly low  (L).    .                         .          --x"  -"....

4.     If the MSA correlation coefficient is <  0.995, qualify the data as estimated (J).

2.12  ICP SERIAL  DILUTION

The serial  dilution determines whether significant physical or chemical interferences exist
due to sample matrix.

2.12.1 Evaluation Criteria

If the analyte concentration  is sufficiently high  (concentration in the  original sample is
minimally a factor of 50 above the IDL), an analysis of a 5-fold dilution must agree within
10 percent Difference (%D) of the original results.

2.72.2 Data Requirements and Retrieval of Data

Forms 1 and 9.
1M-1-D4ORG

-------
                                                  Procedure No.:  M-1-INORG
                                                                   Revision:  2
                                                               Date:  06/30/95
                                                               Page:  22 of 26
2.12.3  Evaluation Procedure
 1.     Recalculate at random the %Ds using die following equation to verify the dilution
       analysis results reported on Form 9.
where:
       I    =  Initial sample, result
       S    =  Serial dUutionresult,instrument reading X 5)
2.     Determine whether thereof negative interference.   Generally ;x a diluted sample
       should offer lesser interference; hence, more accurate and higher reported values.
       However, it is also possible to obtain lower values for the diluted sample, indicating
       a potential for negative interference.  The apparent negative mterference may be
       related to  sample  concentrations  in  the  less  accurate .regions, of  the "IGP
       instrumentation i.e., near the IDL or the upper limit of the linear range. Carefully
       evaluate if the negative interference is real.

2.12.4  Action

1.     Whan the diluted sample results are outside  the range of  90-110% of the original
       sample result, qualify the associated data as estimated (J).

2.     If evidence of negative interference is found, use professional judgement to qualify
       the data.

2.13  FIELD DUPLICATES
                    i •
Field duplicate samples may be taken and analyzed as a indication of overall precision.
These analyses measure both field and lab precision; therefore,  the results may have more
variability than laboratory duplicates which measure only laboratory performance.  It is
also expected that soil duplicate results will have greater variance than water matrices due
to difficulties associated with collection identical field samples.

2.13.1  Acceptance Criteria
                   ji
There are no review criteria for field duplicate analyses comparability.

-------
                                                 Procedure No.:  IM-1-INORG
                                                                 Revision:  2
                                                              Date:  06/30/95
                                                              Page:  23 of 26
2.13.2  Data Requirements and Retrieval of Data

Form 1.
2.13.3  Evaluation Procedure
    •  •    .    •  ••;; •'••^.v.-^i^-.j/.i.-i'i:-,--/.
Samples which are field duplicates should be identified using EPA Sample Traffic Reports
or sample field sheets. The reviewer should compare the results reported for each sample
and calculate the Relative Percent Difference (RPD), if appropriate.     V ;::

2.13.4  Action

Any evaluation of the field duplicates should be provided with the narrative report.  At the
reviewer's discretion,  a table listing the RPDs between the original and.the duplicate
samples may be prepared.

2.14 REPORTING LIMIT VERIFICATION                r

Reporting limit verification is performed to verify that the CRDLs were met; to ensure that
the reported quantitation results were accurate, and to ensure that the ICP data were  not
reported beyond the upper linear range of the instrument.

The positive results themselves are not verified for accuracy, as this would require the  use
of raw data.  However, the undetected values are checked for proper application of dilution
factors and moisture content normalization.  Indirectly, a check on  the reporting limits
would suggest that the sample results have been properly adjusted for the dilution factors
and moisture content.

2.14.1   Acceptance Criteria

Minimally, the  laboratory  should meet  the basic CRDLs specified  in  the  most recent
version of SOW. This implies that the laboratory's IDLs must be equal to or less than  the
CRDLs.  The laboratory has a choice of using the ICP or AA instrumentation for analyses
of metals; however, the sample results for As, Pb, Se and Tl must be greater than 5X HDL.
Otherwise, AA instrumentation must be used.

ICP data  must not be reported beyond the established linear range without sample dilution.
The linear range for each ICP metal is established on a quarterly basis.

-------
                                                 Procedure No.:  IM-1-INORG
                                                                 Revision:  2
                                                              Date:  06/30/95
                                                              Page:  24 of 26
                    I!
Furthermore, the dilution factor and moisture content correction, must be made to the
sample results,  as applicable.  This may apparently raise the reporting limit above the
CRDLs for some analytes.
               . V   i                .
2.14.2  Data  Requirements and Retrieval of Data

Forms 1, 10 and 12.      .

2.14.3  Evaluation Procedure

1.    Verify from Form 10 that the IDLs reported for each analyte and instrument meet
      the CRDLs at a minimum.  There may be several Form 10s for the  ICP, AA, Hg
      and CN analytes.      ,.'•••../•.                           :
                 ป  , ' r~ ••  •  '-•..."''••;-"   - -       '  •    •  - . •"••*•'"_
2..     Verify mat the linear ranges for the ICP metals are reported on Form 12, and no
       undiluted sample data are reported above the linear ranges.       ~    •'

3.     Verify on the Form Is that the CRDLs have been adjusted for any dilution factors.
       and moisture content, as applicable.                      .    '.-,.'.•-.

4.     Verify that the sample results are >SX ICP IDL, if ICP analysis results are used
       for As, H, Se] or Pb.

2.14.4 Action
                   !'       '
If there are any discrepancies found, the laboratory may be contacted by the designated
representative to obtain additional information that could resolve  any differences.   If a
discrepancy remains unresolved, the reviewer may determine qualification of the data is
warranted.

2.15  SAMPLE IปAPERWORK

The purpose for evaluating the sample paperwork is to determine  that the samples  being
validated are indeed  the ones taken  from the site, and have not been tampered  with.
Accurate sample identity is of paramount importance in substantiating  the sample data.
Without unequivocal sample identity and chain-of-custody procedures,  the sample data may
not be  defensible or enforceable.
                  •i
Under  the current CLP contracts, the  original paperwork (i.e., the purge package or the
administrative recdrd) is included  in the data package from the laboratory.  It is assumed
that the data validator is not privy to the original paperwork; therefore, the evaluation
criteria and procedures  described  below apply only to the documents that are generally
1M-1-INORO

-------
                                                  Procedure No.:  IM-MNORG
                                                                  Revision:  2
                                                               Date:  06/30/95
                                                               Page: 25 of 26

included in the data validation package. These documents are the chain-of-custody forms
and Region m Shipping Record.

2.25.1  Acceptance Criteria

Criteria for acceptability or authenticity of the sampling paperwork, document control and
chain-of-custody have been established, by the National Enforcement Investigations Crater
(NEIC), to support the CLP.  Overall criteria are too numerous  and subjective to be
discussed here, but the criteria that apply to data validation are:

       •     The chain-of-custody form  should be properly and completely filled out
             including the sampler signatures, date and time of sampling, sampling station
             identification, analyses requested, traffic numbers,tag numbers, etc. These
             data are minimally required to confirm the authenticity of the sample and its
           •.data,-  . ;-'::CV/^.:; '•''..'  "'             ••  '   •   .'   •". :-:(y  .  :   ' •.
       •     The chain-of-custody must be maintained at all times., The custody, transfers
             between different parties must be signed and dated.

2.15.2  Data Requirements and Retrieval of Data

A copy of the chain-of-custody  form is essential to confirm the identity of the samples.
The Region m Shipping Record is required to identify the field QC samples.  The chain-
of-custody form and Shipping Record are generally located in front of the data package.

2.75.3  Evaluation Procedure

Ensure that the chain-of-custody form was signed and dated by the samplers, and a time
and date were entered for sample collection. The laboratory copy of the chain-of-custody
must have the signature of the laboratory sample custodian. Any errors on the form should
have been  crossed out with a single line through the entry.   Verify that all collected
samples  have unique station identification, traffic numbers and sample  tag numbers.
Ensure that the Region HI Shipping Record correctly reflects the information on the chain-
of-custody.

2.15.4  Action

The action to be  taken in qualifying the data is highly dependent on the  nature of the
problem.  Some  errors in  paperwork are practically  unavoidable in real situations.  An
effort should  be made to reconcile the differences by cross  checking the field notebooks
against the sampling paperwork.  Occasionally, the samplers  may forget to sign the chain-
of-custody; however, the  field  notebooks  may  amply describe the  sampling event.
Problems are  also inevitable in noting or cross-referencing sample tag numbers and traffic
1M-1-INORO

-------
                                                  Procedure No.:  IM-1-INORG
                                                                  Revision:  2
                                                              Date:  06730/95
                                                              Page:  26 of 26

numbers.  Generally, there are several alternate sources of information to substantiate or
refute the problem.

Any discrepancies found in the paperwork must be immediately brought to the attention of
the EPA RPM or SPO.  Clearly define the problems in a memorandum to the responsible
parties.  Attach marked copies of the chain-of-custody forms to substantiate the findings.
1M-1-INORO

-------
                                    Case No.;  .
                                    SDG No.:Z
                                    Data Reviewer/Date:,
                                    Site:-  --1 •••:.•'• :--' >'S'.
                                                                                    Procedure  No.:   1M-1-INORG
                                                                                                   Revision:   2
                                                                                             Date:   06/30/1995
                                          "'••"-':-' r •- ••'   •'•
         Sample
                                                                        Hg (28 Days)
                                                                                         CN (14 Days)
Number
Date
Coll.
Date
Reed.
Lab.-

                                                       ปS
: Days
Qual.
Date
Prep.
Anal.
Hold.
Time,
Days
Qual.
Date
Prep.
Anal.
Hold.
Time,
Days
Qual.
            II
        ••//*


            1 1
        77-


                             ••••• •
                    II
                    11  -
                    / A
                    7 /
                    1 /

                    / 7  :-
                                                     :••/'
         //
         /  /
         /  /
                              •  * -.

-------
                      Case No.:.
                      SDG No.:
                  Data Reviewer/Date:.
                      Site:
 Procedure No.:  IM-1-INORG
                Revision:   2
__         Date:  06/30/1995
Table  Dt-1-INORG-CAL.    TAL Metals and Cyanide Calibration and Blank Qualifier Summary
ICP Instrument ID:
Instrument Calibrated? __Yes _Nb
ICP Date Started:
ICP Date Ended.'
AA Instrument IDs: As P
Instruments Calibrated?
AA Date Started: ' . v
AA Date Ended: ';.
Analyte (Criteria)
Aluftiiinifii • • Al *
-Antimony , Sb
Arsenic . •' >'.As
Barium Ba
Beryllium Be
Cadmium Cd
Calcium • Ca
Chromium Cr
Cobalt Co
Copper Cu
Iron Fe
Lead Pb
Magnesium Mg
Manganese Mn
Mercury . Hg
Nickel Ni
Potassium K
Selenium Se
Silver Ag
Sodium Na
Thallium Tl
Vanadium V
Zinc Zn
Cyanide CN
,V ".'.' "' •'•' ." '^f'-"'.
v" -'"''•" -."^"i

!.9&8i$i
'•'".'• '•"'?' .'?'•• *V
•': ;:';-'^i
•'r;"-..'?sf::4



















!
'.
ฑfr
•fat*
>^3
1
..?'•>ป
Sample Identifier
1. , •
2. .'••••• A ' ' . ' •
'•3;'... -:.'.:'.ฃ :*- " • • ' ' . •• • '
,:4i%^fe;:'. '.. ;• '• : . ••'.;• •-. -.
\5&3i!i&1 ••'•• •••• -••' '. •"•• ' • - •;
^6;^^^'^- '••'•:• •"••• ' ' '->••'- '•'.'
^j^i^ซ^:'-;---:': ••'• ' •'..'..:'••'•••-.
^B'^^l^-'t'-.''^''.'.-'' '.: '" •..'....-.'• .•-.•':
W$f$$&$$&" •'• . . •••'?•• '••' "••...•
-•flO^JtSl^- ""-•'• ' ' : ' • '.'--.'
• •.ป>&ซ.1i>



il$8$
vSIHB
-i-'ปS •' W"




-















s;ซS53<ซ •.;•• • •' • • • . ••...•
AgJWf>r-.V.-. ;

WS:^:
^p?
SSI''--"
^Ix!"
H'l- '
'*-U*-r '.' .. -.



,















:% •:••-•'.,
Qual.
•'.•.'-%'
'. /:••""_


. ' .






•












&&K
•••••?' ••''' ••'•:• ••>••..•-...'.
18. ;:' '••- • .•••.-•• ' • • • . •
-.19..- • .' •' .-••.•• . ••
'20.', • •••:'••'- '. '•

;^5
'';I "•

. . ••• • '..




















g&^
*lXitiL:
• ., r '.•'*• '
'::'-'' '• '-: V
>* • ' ":


^


















.is??'.
•'xn)L:
•^•'•'. -.
.;;-..'' •



--.















-


Qual.

^






















• The ICV and CCV criteria: He = 80-120% R, CN = 85-1 15% R.

-------
    Uase No.:.
   SDG No.: ~
Data Reviewer/Date:.
    Site:
/
 Procedure No.:  IM-1-INORG
              Revision:   2
	     Date:   06/30/1995
Table IM-1-INORG-SPK. TAL Metals and Cyanide Matrix Spike, Duplicate and ICP Serial Dilution Qualily Control Summary
Analyte
(Criteria):
Alnabnm AI
*afiซr.Wป~
•jiaa&&$&
jtarhim-^Ba-'
BerfOhan •-• : fie
Cadmium ' '.Cd
ciiM^t*."
'titaoUn^Ck'-
:0*il$fฃcoV
Copper >-•-< 'Ob •
'tiriB^v'W*;
Lead."-:.''- :"''."Po
Majmeffam ME
MangBncff Mn
Mercury. HE
N34yป/ ' N! '
Potassium ' 1C
Selenium Se
Silver Ag
Sodium Na
Thallium Tl
Vanadium V
Zinc Zn
Cyanide ' CN
Matrix Spike/Duplicate
Rec,%R
75-125
• ...
' >
•i ": ; ' •
-. : -••''.

•. ' :•
'; ;• •• •.:. .'''"
'.•• , '" ••."•.-.
"o. :•'•:"'.'••'
•; .i '•• " . ' ' •

' •;-'-• •,'...'




-







RPD
25(35)ป

•'•''"• •.'-.•• ••'•'•
'.""•'•uh-:
• ••••.: ;••.•.•'

~. *' . :
/••••••:•::,•;,
•;;..-.;..,..::•;.••
''./•.:•'.,:•.;•'.••
' " " ";. - •'.'
-• ; :: •'';•-.'''
.'• ";'•'• ' '


_ '- • . •._









• Qualc,
".', v • '• . -
H^|^::
•;;sf^'^.
•;W*&



:?^g^i
'.'^'S??*?'"'-"
;g^^:
"•.ป-*: ->",^ "- -..' ;'
/r-.^pf"
v.';^*;;v.
•••.:•••-•".:ป
. " •- • .
•'• . •
'







ICPSer. Dil. .
+/-1096
'• .'•-'•' • '•'
-ฃ^-." '
"••l^"^-"-
r^^v ..;
~-:^/'<:-'


•fe;^:r- '
^^•tB::',:;
i^v-y -;
--. ~. '**-•-. .
"!•' '. 7'* -?Hi' .- -; . •
: ._ .-'. •-.. .""











Quals.


'_ .. •




f.jฃ
















Post Digest/Anal. Spike*,
ICP
75-125%







'r •*' • •






-• . -









AA
85-115%
•' ;. :.
•:' ' ' • '^
'..-.- .."'ii
' :' -:v'-
.'.•' ' '•":"
•;.
- . . V.
.., . .:.^-iซ.


" ' •' '.













Quals.
. jtii •' :iX:.',*>-t':







$$%$!$:
^SS
"Si^v
•'^f^
-•' *-*• '.- — - -'
•'• :'.'.-"'''J-.-:-:
-• """ "^-Z..
•'••'-^•^'








LCSRecoverv
80-120%-
: ?:,-; .

$$&&ฃ•
H^':V:"'
r?is^--.-
:^r-;v •
iR'-'v-
^fe:v,
^t^:^ :•:•;
S^P.V :t\
''i&?--'" '••.:
**-iV'!'i. •'•".
.:...: •••' • ••
•' ' ' . . *


• . •
t

'




Quals.
1 •
l'>'V;:- • ••
'•'•: •^.'••••'


.: • •* •
-• '-. r . ' .

•••:'•-':• . - '• :
.'. "."." •SJr>"
' "v

•










. Note: See the SOP for the exact criteria under special situations.
•TThere are no criteria at the present time to qualify the ICP metals or Cyanide based on the post-digestion spike recoveries.
* The precision RPD for soil samples is 35. . '
'ffhe-range of 80-120% R is used only for the aqueous LCSs. Solid LCS samples have specific recovery ranges for each analyte.

-------
        Case No.:.
  SDG No.:	
Data Reviewer/Date:.
       Site:	
Procedure No.:  Hf-1-IHORG
               Revision:  2
         Date:  06/30/1995
Table IM-1-INORG-<|L. Summary of Data Qualifiers ,
Sample
Number

- --. - _-. ,_


















Analyte i .
Al















- •




Sb




















As




















Ba












-







Be




















Cd



















•-•
Ca














'. .





Cr





"l
•(




\








Co










-
/


. *





Cu

, .-












. ' ; ''•'/
' -ill-

• ••


Fe


, w • . •
<'• ,•*'
*Jf
;.r; •:.-.'. \-:-



•*




-.. ..•
':'.' ;v. •
• • • . .
•i '• '
. .

.Pb
•-

'•&•
:iOj-
.•^-..C?*ii>i
VW0^
•;• ';,:.••.''!;

i .



._'

. . t >:' '
'• :-:d-':-
• • ' ซ )

t
\

;Mg.

' *.
*=- ••,-'•
:,V-'4t
.j!,v,.'V^:O
^
Wf&
tftx-




"•



'•f?&
""* .-. ., '



'. •
Mn

,'-•.--''.•
•ป•• f.'*
it-vSV-f.
iH
IP
1ฎ
:>'";•••






v.'V'^'-''
!f^;
••''. tv.-' •




Hg

'"":•" .
^;v
^-•.A'!*-
iK
K>XHW.
^
IK


h."
'ป.
N.
-


vV?.";!:'.
utr. •.'.'•-':•:
s*T.ซcs ;
^^••:
•:*irt*-
^y'4''

-


Ni

; . . t
V,'."'.V.-"
i/vte
^^K
ซ
t?.^f-
t^;
:'.-'.. K








.-.
'•A-"":{' •••
.if.;-:;''




K

,-.'.. -i.
•^
ijjj^
^
i;=j-'*ii
•';1;SJ;y


. " .-.





•'• ;'-U
.•..t1-*
.' '.<<'"&•
-. 4' - *>
-'. .-Iv'-V.
T ' '



,Se
>
'v •*...' ;
• ;.v-'-.'--
^
SI
^
•v?>3;ฃ'-




*


.•".. '.
:i- 1:5>'
•-.- .. -. 5-
k**'.'*;
;*ฃ*)-
f^-s-1'
•;..-."}•

• ' '

Ag


- ' -
'<-*'.' "• .
.".' : .
-it : '/
*%*• ••:
•\' •-' •
_;!:X. '•'•.




-.



',? ..
*.




Na











ซ








Tl




















V





t .














Zn





-













it
CN




•
















-------
                         Case No.:_
                         SDG No.:"
                     Data Reviewer/Date:,
                         Site:	
                                                                        rocedure No:   IM-1-INORG
                                                                                     Revision:   2
                                                                        —    Date:   06/30/1995
                                    Table IM-1-INORG-IRDA
                                          EPA Region m
                                 Inorganic Regional Data Assessment
Case No.
                                              Site
Laboratory
SDG No. _
SOW No.•
IPO: Action
                        FYI.
 Number of Samples/Matrix.
 Reviewer (if not CRL)	
 Reviewer's Name  .  •
. Completion Date	.	
Holding Times
Calibrations .. '
 ds
                                     Data Assessment Summary
                                          ICP    .    AA
                                                                                 Cyanide
Duplicate Analysis    •
Matrix Spike       .
Post Digestion/Analytical Spike
MSA
Serial Dilution
Held Duplicates  .
. Reporting Limit Verification
Sample Paperwork
      o  =
     M  =
      Z  =
      X  =
Action Items:
              Data had no problems/or qualified due to minor problems.
              Data qualified due to major problems.
              Data unacceptable.
              Problems, but do not affect data.
Areas of Concern:
Notable Performance:

-------