United States
Environmental Protection
Agency
Industrial Environmental Research
Laboratory
Research Triangle Park NC 2771 1
EPA-600/7-79-207
August 1979
Evaluation of Sensitized
Fluorescence for
Polynuclear Aromatic
Hydrocarbon Detection

Interagency
Energy/Environment
R&D Program Report

-------
                 RESEARCH REPORTING SERIES


Research reports of the Office of Research and Development, U.S. Environmental
Protection Agency, have been grouped into nine series These nine broad cate-
gories were established to facilitate further development and application of en-
vironmental technology. Elimination of traditional grouping was  consciously
planned to foster technology transfer and a maximum interface in related fields.
The nine series are:

    1. Environmental Health Effects Research

    2. Environmental Protection Technology

    3. Ecological Research

    4. Environmental Monitoring

    5. Socioeconomic  Environmental Studies

    6. Scientific and Technical Assessment Reports (STAR)

     7. Interagency Energy-Environment Research and Development

     8. "Special" Reports

     9. Miscellaneous Reports

  This report has been assigned to the INTERAGENCY ENERGY-ENVIRONMENT
  RESEARCH AND DEVELOPMENT series.  Reports m this series result from the
  effort funded  under the 17-agency Federal Energy/Environment Research and
  Development Program. These studies relate to EPA's mission to protect the public
  health and welfare from adverse effects of pollutants associated with energy sys-
  tems. The goal of the Program is to assure the rapid development of domestic
  energy  supplies in an  environmentally-compatible manner by providing the nec-
  essary environmental  data and control technology. Investigations include analy-
  ses of the transport of energy-related pollutants and their health and  ecological
  effects;  assessments  of, and development of,  control technologies for energy
  systems; and  integrated assessments of a wide-range of energy-related environ-
  mental  issues.
                         EPA REVIEW NOTICE
  This report has been reviewed by the participating Federal Agencies, and approved
  for publication. Approval does not signify that the contents necessarily reflect
  the views and policies of the Government, nor does mention of trade names or
  commercial products constitute endorsement or recommendation for use.

  This document is available to the public through the National Technical Informa-
  tion Service, Springfield, Virginia 22161.

-------
                                          EPA-600/7-79-207

                                                August 1979
Evaluation of Sensitized  Fluorescence for
     Polynuclear Aromatic Hydrocarbon
                       Detection
                             by

                          T. R. Smith

                 TRW Defence and  Space Systems Group
                         One Space Park
                    Redondo Beach, California 90278
                      Contract No. 68-02-2689
                           T. D. 104
                     Program Element No. INE624
                  EPA Project Officer: Larry 0. Johnson

                Industrial Environmental Research Laboratory
                  Office of Energy, Minerals, and Industry
                   Research Triangle Park, NC 27711
                          Prepared for

                U.S. ENVIRONMENTAL PROTECTION AGENCY
                  Office of Research and Development
                      Washington, DC 20460

-------
                                 CONTENTS

Abstract	    1v
Tables	      v
Figures  	    v1
1.0  INTRODUCTION  	      1
2.0  EXPERIMENTAL CONDITIONS 	      4
3.0  SPOT TEST PROCEDURES	      6
     3.1   Results	      6
     3.2  Procedures	      6
4.0  DETECTION LIMIT AND SAMPLE ANALYSIS 	      8
     4.1   Results  	      9
     4.2  Interferences  	      9
     4.3  Methods of Identification by GC/MS 	    12
     4.4  Combustion Effluents 	    13
     4.5  Coke Oven Effluent Samples	    13
     4.6  Spot Test Versus GC/MS: Detection Limit  	    14
5.0  SUGGESTIONS FOR FUTURE STUDY  	    32
6.0  CONCLUSION	    37
7.0  REFERENCE	    38
                                    111

-------
                                 Abstract

     A flourescent spot test used for detecting the presence of polynuclear
aromatic hydrocarbons (PAHs) has been evaluated as a screening technique
for samples to be analyzed by gas chromatography/mass spectrometry  (GC/MS).
The spot test is based on the phenomenon of sensitized fluorescence and
is capable of easily detecting 100 picograms of PAH in a 1  microliter
sample, a level of sensitivity adequate for the screening of combustion
effluent samples.
     Two interferences were observed:  1)  Samples which are highly  colored
required dilution to allow viewing of the  fluorescence level and 2) samples
containing substantial amounts of phthalate esters produced false positive
results.  No false negative results were observed in this study.
     It is our conclusion that the spot test procedure is adequate  for the
screening of combustion effluent samples prior to GC/MS analysis.
                                     1v

-------
                                 Tables
1.   Results  of spot test  analysis of standards	   10
2.   Results  of spot test  analysis of samples from
     industrial  sites 	   11

-------
                                   Figures

 1.   Total  Ion chromatogram of a  combustion effluent sample ....    15
 2.   Mass chromatogram of m/e  252	    16
 3.   Mass chromatogram of m/e  202	    17
 4.   Total  1on chromatogram of a  typical coke oven sample  	    18
 5.   Mass chromatogram of m/e  202	    19
 6.   Mass spectrum -  scan number  454	    20
 7.   Mass spectrum -  scan number  476	    21
 8.   Mass spectrum -  scan number  738	    22
 9.   Mass chromatogram of m/e  252	    23
10.   Mass spectrum -  scan number  710	    24
11.   Total  ion chromatogram of detection limit  test sample  ....    25
12.   Mass chromatogram of m/e  202	    26
13.   Mass chromatogram of m/e  252	    27
14.   Mass chromatogram of m/e  300	    28
15.   Mass chromatogram of m/e  302	    29
16.   Mass chromatogram of m/e  202	    30
17.   Mass chromatogram of m/e  252	    31
18.   HPLC of benzene, chloronaphthalene and benzo(a)pyrene  ....    34
19.   HPLC of naphthalene, fluorene, phenanthrene
      and benzo(a)pyrene	    35
20.   HPLC of coke oven sample	    36
                                     v1

-------
                             1.0   INTRODUCTION

     This document represents the final  report on the evaluation of the
Polynuclear Aromatic Hydrocarbon Spot Test.   This report is submitted in
partial fulfillment of the requirements  for EPA contract No.  68-02-2689,
Technical Directive 104.  The work was conducted by TRW under the direction
of EPA Task and Project Officer, Dr. L.  D. Johnson.
     The original fluorescence spot test discussed in this report was
developed by Arthur D. Little, Inc.  under contract to the EPA and is based
on the phenomenon of sensitized fluorescence.   The objective  of this effort
was to determine if the test could be used as  a screening aid for samples
submitted for PAH analysis by gas chromatography/mass spectrometry (GC/MS).
The procedure can potentially provide a  low cost screening technique to
predetermine the presence of PAHs.
     This test was evaluated specifically for its detection of PAHs.  The
importance of PAHs as a class originate  from the serious health hazard
posed by certain components of this class of compounds.  Specifically,
benzo(a)pyrene and dibenzo(a,i)pyrene are compounds within this class
which have been shown to be potent cancer causing agents.  The potential
danger of contracting cancer is incurred when these compounds are inhaled
into the lungs or come into contact with the skin.  Therefore, it is
important that these compounds be detected and identified in effluents
which could come into contact with the population.
     The analysis schemes thus far implemented for the determination of
PAHs encompass several analytical approaches including nonspecific
analysis, detailed quantitative and qualitative analysis, and specific
analysis for a single PAH.  The limits of detection have been reported to
be at the 10 picogram level in some instances.  The analysis  times range
from 15 minutes to several hours with the evaluation if the test
results occupying an equal amount of time.  The PAH spot test is a survey
technique which provides information as  to the presence or absence of PAH
as a class and no specificity is provided.  The level of detection is in
the picogram range and the analysis time is on the order of 15
minutes.  Gas chromatography/mass spectrometry can provide qualitative
and quantitative information on the PAHs present in an effluent.  Sample

                                     1

-------
analysis time is typically between 1  to 2 hours with the data
evaluation consuming at least an equal  amount of time.   The limits  of
detection for the GC/MS analysis normally approaches 5  nanograms  per
PAH.
     For general purpose determination  of PAH in complex mixtures the use
of GC/rlS is unexcelled.  The gas chromatograph is used  to separate  the
complex mixture into individual  component peaks and the mass spectrometer
is used as a detector for both qualitative identification of the  PAH and
for quantitative determination.   For  simple cases where only a few  PAH
are present, the use of packed gas chromatographic columns 1s  recommended.
Packed columns are generally easier to  use, require a shorter  time  for
analysis and are cheaper than capillary columns.  When  highly  complex mixtures
of PAH are present or significant interferences occur,  the use of capillary
columns becomes necessary.  Chromatographic column selection is usually
dependent on the volatility and polarity of the PAH to  be determined.
     The major limitation of the gas  chromatograph is the volatility of
compounds to be analyzed.  In order to  be determined by GC/MS, PAHs must
produce good chromatographic peaks at temperatures below 300°C.  At or
above 300°C even the most stable GC column liquid phases produce high
background in the mass spectrometer.   This GC temperature limitation means
that high molecular weight PAHs (above  molecular weight 300) cannot be
reliably determined by this technique.   For this reason some PAHs which
produce fluorescence in the spot test may not be analyzable by GC/MS.
     As discussed above, the,analysis of samples by GC/MS for the routine
detection of PAHs expends costly GC/MS  instrument time  as well as the time
needed to interpret the data.  The PAH  spot test is proposed to serve
as a screening tool for the GC/MS analyst.  This pretest will  serve as a
go/no go analysis.  That is, if the test indicates the presence of PAH,
the sample will be run via GC/MS.  If no fluorescence is observed it will
not require analysis.  Since the PAH spot test requires only about 15
minutes of effort and no instrument time, implementation of this test
should result in a significant decrease in PAH analysis time and cost.
Clearly, this is contingent upon several factors.  First, for maximum
benefit to the combustion environmental assessment program  sensitivity
should equal or exceed that of GC/MS for other  projects the sensitivity

-------
requirement varies with the job.   Second,  interferences  which  would
result in false negative findings must be  either  identified  and  taken
into consideration or they must be nonexistent.   The  experiments  reported
here were designed to evaluate these possibilities.

-------
                       2.0  EXPERIMENTAL CONDITIONS
     Phenanthrene and benzo(a)pyrene were chosen as model  analytes  to test
the spot test procedure.  This choice was based on the facts that:
          t   Phenanthrene and benzo(a)pyrene are representative of PAHs
              which are likely to occur in combustion samples.
          •   The range of aromatic fused ring compounds detectable by
              GC/MS is covered by these two compounds.
          •   Any difference in sensitivity of the spot test to PAHs should
              be discernable by the comparison of the response of these
              two components,
          »   and lastly, because of benzo(a)pyrene's carcinogenicity, it
              is critical that the presence of this compound be detected
              by the spot test at low concentrations.
     An ultraviolet Chromatovue Cabinet, Model C-70, was used for ultraviolet
exposure of the samples.  This unit contains both 254 and 365 nm lamps of
which the 254 nm source is used for the naphthalene-PAH fluorescent
detections.  Whatman 142 ashless filter paper was used as a substrate for
the sample analysis.
     The phenanthrene and benzo(a)pyrene used for the prelimenary study
were obtained from Analabs, Inc., North Haven, Conn.  The solvent, methylene
chloride, v/as obtained from Burdick and Jackson, Inc.,  MusKegon, MI.
The sensitizer, solution, naphthalene was prepared at a concentration of
60 yg/yfc 1n methylene chloride.
     The gas chromatographic/mass spectrometric analyses were done on a
duPont model 321 GC/MS.  The key operating parameters of the GC/P1S were
as follows:

        Column  -  2 m. X 3mm ID glass packed with 3% OV-101 on 100-120
                   Chromosorb WHP
        Carrier Gas  -  Zero grade helium at 30 ml/min
        Oven Temperature  -  100°C to 295°C programmed at 8°C/min
        Injector Temperature  -  300°C
        Transfer lines and separator Temperature  -   270°C
        Mass Range  -  40-450 AMU
        Scan Function up - 1.9 seconds
                      down - 0.0 seconds
                                     4

-------
                      hold at top  -  0.0 seconds
                      hold at bottom -  0.1  seconds
     The  high  performance  liquid chromatograph (HPLC)  used for this report
was a duPont model  850 equipped with a  variable wavelength ultraviolet
spectrophotometer.   The wavelength monitored on all  chromatograms  was
254 nm.   For all  separations  a Zorbax-ODS,  4.6 mm ID X 25 cm,  liquid
chromatographic  corumn was used.   All solvents were  obtained from  Burdick
and Jackson, Inc.
     The  samples  used  to determine applicability of  the spot test  procedure
to combustion  effluent samples were  provided by ongoing EPA sponsored
projects.   Specific sample type identifications are  presented  in the results
section of  this  report.

-------
                         3.0   SPOT TEST PROCEDURE
3.1   PREPARATION  OF  SAMPLES

     No  preparation  of  the sample  is required.  The sample should be in a

solution to  allow easy  application  to the test substrate.  Solids and gases
can  be analyzed with some modification of the application technique


3.2   PROCEDURE

     A Chromatovue ultraviolet cabinet Model C-70 was used to expose the

samples.   The 254 nm lamp source was used.  A 7.0 centimeter circular
ashless  Whatman #42  filter was used as the support for the samples.  The

sensitizer used was  naphthalene at  a .concentration of 60 yg/pJi in methylene
chloride.

     The spot test is carried out as follows:

        1.   Draw  three  circles in pencil on a piece of Whatman #42
            filter paper as close to one another as feasible.

        2.   Using a  10  y£ syringe,  spot the filter paper with the
            sensitizer  (naphthalene).  1 mlcrollter of sensitizer
            should be spotted on the leftmost circle and 1 microliter
            of sensitizer should be spotted on the center circle.
            The spots should not be allowed to overlap.  The
          .substrate should be supported so as not to touch any
            surface.

        3.   Allow the solvent to evaporate from the filter paper.

        4.   1 microliter of sample  is then applied to the rightmost
            and center  circles.  Again allow the solvent to evaporate
            and do not  allow the substrate to touch any surface
            until the solvent has evaporated.

        5.   The filter  paper is placed in the Chromatovue cabinet.

        6.   Compare  the spots visually with the unaided eye.

        7.   If the center spot (sample plus sensitizer) fluoresces
            brighter than the sensitizer spot (leftmost), the presence
            of PAHs  is  suspected.   Also, any significant difference in
            color between the sensitizer only spot versus sensitizer/
            sample spot is also considered a positive indicator for
            PAHs.

        8.   Samples which give positive results by the spot test should
            be submitted for GC/MS  analysis to identify the source of

-------
the fluorescence.

-------
                 4.0  DETECTION LIMIT AND SAMPLE  ANALYSIS

     In order to determine the level  of detection attainable with the
equipment available, a set of standards were prepared  and  tested.
Phenanthrene and benzo(a)pyrene were  chosen  as  the test  standards.  As
previously discussed, phenanthrene and benzo(a)pyrene  were chosen because
both are considered representative of the PAHs  likely  tc occur  in
combustion effluent samples.   Their differences in structure may also cause
differences in their detectability by the PAH spot test.  Additionally,
the range of aromatic fused ring compounds detectable  by GC/MS  is covered
by these two compounds and the presence of benzo(a)pyrene, because  of its
carcinogenicity, must be detectable by the spot test at  low concentrations.
     The levels of standards  prepared and the results  of the analyses are
presented in Table 1.  The limit of detection of the spot  test  is
extropolated to be approximately 100  plcograms  for these two materials.
That is, a 1 micro!Her sample containing 100 picograms  of PAH  is
easily detected.  This observation differs by an order of  magnitude
from the detection limit of 10 picograms previously reported  .   The
difference in the reported detection  limit is probably due to  the point
at which the investigator decides to  declare a  significantly distinguishable
difference in the fluorescent intensity of the  sensitizer  alone versus  the
sensitizer/sample spot.  Since this decision is subjective it  was decided
to choose a level that was easily discernable by personnel not familiar
with the test.  A lower detection limit is achievable  but  it would  require
the operator to make judgements based on subtleties of intensity and  hue
and would require extensive operator training and experience.
     Another factor which must be taken into consideration when detection
limits are compared is the ultraviolet lamp intensity.  Any difference  in
lamp intensity will result in a difference of fluorescence intensity  and
consequently, the detectable difference between the sensitizer only and
the sensitizer/sample spots would be concommitantly effected.   It is  an
accepted fact that lamp intensity varies with age and  it is not unusual
to encounter significant differences in lamp intensities in an apparatus
such as the Chromatovue C-70.  Differences in lamp intensity as a source of
variation of the detection limit was not investigated as a part of this
task.
                                     8

-------
4.1  RESULTS
     A series of industrial  samples  which  gave  negative  results  for  both
the PAH spot test and GC/MS  analysis were  spiked  with  a  known  amount of
benzo(a)pyrene (100 pg/v«.) and retested by the  spot  test.   As  expected, the
samples after spiking yielded a positive indication  by the  PAH spot  test.
     Over 20 samples from industrial sites have been tested by both
GC/MS and the fluorescence spot test.  A partial  listing of these  samples
and their results are given  in Table 2.  Approximately 10%  of  the  samples
which gave a positive result of the  spot test showed no  PAHs by  GC/MS.  As
will be illustrated below, the spot  test has  a  level of  detection  which is
lower than that of GC/MS, when the GC/MS is used  in  the  traditional  scanning
mode.  Taking this fact into consideration, 1t  is not unreasonable to
expect a small percentage of the samples to prove positive  by  the  spot  test
and negative by GC/MS.  But  what is  more important is  the fact that  in  all
cases where the spot test gave negative results the  GC/MS also found no
PAHs.  That is, no false negatives were encountered.

4.2  INTERFERENCES
     The effectiveness of the PAH spot test is  lessened  when a highly
colored sample is encountered.  If the sample is  opaque  or highly  colored,
the sample must be diluted.   This dilution is to  allow the operator  to
view the fluorescence of the sample  without interference from the  sample's
color.  The detection limit of the spot test is increased proportionatly
as the sample is diluted.
     An example of this interference is demonstrated by  the combustion
effluent samples studied.  It is not unusual  for combustion effluent
samples to posses a deep yellow color.  When tested, the yellow color of
the sample combined with the blue color of the naphthalene sensitizer
forms a purple-blue sensitizer/sample spot.  This purple-blue color of the
sensitizer/sample spot is different  in color from t'ne blue sensitizer only
spot.  Since it has been noted that  in some cases a difference  in hue is
an indication of PAH presence, this   interference may  be the source  of
some confusion and result in unconfirmed positives.
     When the sample is highly colored, it must be  diluted so that  the color
does not interfere with the fluorescence.  For samples where  color  was a
                                     9

-------
                                  Table  1
                Results of spot test analysis  of standards
      Phenanthrene
                                 Benzo(a)pyrene
Amount
 80 ng
  8 ng
160 pg
 80 pg
Test Results
 Positive
 Positive
 Positive
 Negati ve
Amount
490 ng
 49 ng
147 pg
 98 pg
 49 pg
Test Results
 Positive
 Positive
 Positive
 Negative
 Negative
                                     10

-------
                            Table  2
Results of spot test analysis  of samples  from  Industrial  sites
                      Coke Oven Extracts
                           Spot Test             GC/MS
Sample
  A
  B
                            Positive
                            Positive
Positive
Positive
    A
    B
    C
                Combustion Effluent Samples
                  (Commercial  Oil  Burner)
                          Negative            Negative
                          Positive            Positive
                          Positive            Positive
     A
     B
     C
     D
     E
     F
     G
     H
                Combustion Effluent Samples
                 (Coal Fired Power Plant)
                          Negative            Negative
                          Negative            Negative
                          Positive            Negative
                          Negative            Negative
                          Negative            Negative
                          Negative            Negative
                          Negative            Negative
                          Positive            Negative
                               11

-------
source of interference,  the dilution  was  achieved  by  adding more sensitizer
to both the sensitizer/sample and sensitizer only  spots.   The  sensitizer was
added equally to both spots until  the sample color no longer interfered.
This was done to maintain the total  amount of sample  constant.  As  of  this
writting, no more than 3 m1crol1ters  of sensitizer has  been required.
For samples that are highly colored,  it is also  suggested  that the  samples
not only be examined immediately after application to the  filter paper,
but also after the sample has been exposed to the  UV  source for a total of
5 minutes.  This delay allows for a clearer distinction between the
sensitizer and sensitlzer/sample spot. During this 5 minute period the
fluorescence of the naphthalene diminishes and if  there are PAHs present
in the sample, the rate of fluorescence decay will be more rapid in the
sensitizer only spot.  Another suggested  technique is to also  view  the
sample under the 365 nm source.  In some  cases this has been found  to
enhance the differences in the spots.

4.3  METHODS OF IDENTIFICATION BY GC/MS
     When using GC/f1S polynuclear aromatic hydrocarbons are  best  identified
as isomer classes based on molecular weight.  Once a  chromatographic
separation has been achieved and the large quantity of mass  spectral data
is stored in a computer based system, individual mass chromatograms for
specific molecular weights can be displayed.  After it is established
which classes of PAH compounds are present in a given sample,  standards
of the individual components or relative  retention time data  can  be used  to
identify specific isomers.
     PAHs typically show very strong molecular ions  (the mass  representative
of molecular weight) in their mass spectra because of the stability of their
aromatic ring system.  This strong molecular ion  allows for a very simple
identification of molecular weight even in complex mixtures.   The
disadvantage of this strong molecular ion production is that isomeric
PAHs produce spectra which are virtually identical.  The mass  spectrometer
is a rather poor device for distinguishing  isomers since once  ionized a
given compound will seek its most thermodynamically  stable form.   Since
isomeric PAHs produce similar or identical mass spectra,  the  determination
of structure is best accomplished by comparision  of  their GC  retention
times to standards.  That  is, for qualitative identification  of specific
                                     12

-------
isomers, standard addition of known  PAH  compounds  is made  to  each  sample.
An increase in GC peak intensity without broadening  is  used as  confirmation
of identification.   Once a single isomer of a  given molecular weight  is
identified, relative retention time  data can be  accurately applied for
identification of isomers where standards are  not  available.

4.4  COMBUSTION EFFLUENTS (COAL BURNING  PO'-JER  PLANT)
     Figure 1  is an example of the total ion chromatogram  of  a  combustion
effluent sample which was analyzed by both GC/MS and  the spot test.   Both
the GC/MS and  PAH spot test yielded  negative results.   Examination of the
mass chromatogram for m/e 252 (Figure 2) indicates that benzo(a)pyrene,
perylene, etc. were not detected by  GC/MS.  Also,  the  mass chromatogram
for m/e 202 (Figure 3) shows no indication of  fluoranthene, pyrene, etc.
This technique of using selected mass values to  search GC/MS  data  for PAH
is used routinely in our laboratory.  By using mass  chromatograms, GC/MS
data can be rapidly searched for the most common PAHs.  This  technique is
less time consuming than displaying the mass spectra for all  of the peaks
observed in a  GC/MS run.
     The data  displayed in Figures 1-3 are typical of samples run  from this
combustion source.   Some samples of the series gave positive  results on  the
spot test but  all proved to be negative by GC/MS.

4.5  COKE OVEN EFFLUENT SAMPLES
     When coal is pyrolyzed by a burning or coking process,  significant
amounts of PAHs are produced.  In the burning  process most of the organic
material is destroyed through oxidation, producing water and carbon
dioxide.  In  the coking process, however, significant amounts of organics
are evolved and emitted to the atmosphere.  For these reasons the coke oven
samples were  an essential inclusion into the experimental  scheme.
     Figure 4 is a GC/MS  total ion chromatogram of a typical  coke oven
sample.  This  sample was  shown to contain PAHs by the spot test.  Presented
in  Figure 5 is  the mass chromatogram for  the m/e  202 of this sample.  In
this figure,  peaks containing m/e of 202 are evident.  The possible
identities of these peaks include the PAHs, fluoranthene  and pyrene.  An
examination of the mass spectra of these peaks which are  shown in Figure  6,

                                     13

-------
7, and 8 will more clearly aid in defining  their  identity.   These mass
spectra confirm the presence of pyrene and  fluoranthene.   The mass
chromatogram of the 252 ion is illustrated  in Figure  9.   Two peaks  are
clearly shown which have significant abundance at m/e 252.   Since the
benzopyrene isomers have a molecular weight of 252 and elute in  this
region of the chromatogram, they serve as likely  candidates.  Examination,
of the mass spectrum of the peak eluting at scan  710  (Figure 10) confirms
the presence of at least one benzopyrene isomer.

4.6  SPOT TEST VERSUS GC/MS:  DETECTION LIMIT
     Figure 11 is the GC/MS total ion chromatogram of a sample which  gave
a strongly positive PAH spot test result.  The amount of sample  injected
onto the GC column was 3 m1crol1ters.  This is 3  times the amount
of sample used for the spot test.  Figures  12 through 15 are the mass
chromatograms for the m/e's 202, 252, 300,  and 302, respectively.   These
represent the ions which are indicative of the PAHs typically  present in
these samples.  In Figure 16, the area of the m/e 202 peak maximizing at
scan 447 (15:00 minutes) is 53K counts.  Extrapolation of this number to
an area still distinguisable in the mass chromatograms suggests  that the
sample size could be reduced from 3.0 fefc to 0.3 y£, an effective dilution
of 10, and still be detectable by GC/MS.  Figure  17 is the mass chromatogram
of the 252 ion including the areas calculated for the components containing
this ion.  The previously noted extrapolation is also applicable in this
case.  For the purposes of comparison, this will  be considered the limit of
detection for the GC/MS analysis.  That  is, the  sample, if  diluted by more
than a factor of 10, would not contain an amount of  PAH that could clearly
be detected by GC/MS.
     In order to compare the aforementioned results  to the  detection limits
of the PAH spot test,  the sample was  diluted  and the  resultant dilutions
subjected to  the spot  test.  The sample  when  diluted  by a factor of 100
still gave a  positive  result by  the  spot test.   The  subsequent level of
PAHs in the  sample after a one hundredfold dilution  was previously
extrapolated  to be below the detection  limit  of  the  GC/MS.  Consequently,
using this sample  as an example, the  limit of detection of  the PAH spot
test has been demonstrated  to  be more than adequate  for  the purpose of
determining  which  samples  are  to be  further  analyzed by  GC/MS for  PAHs.
                                     14

-------
tea. a
 BIC.
 Figure  1.   Total  ion chromatogram of a combustion  effluent sample.

-------
Cr>
                                                                                               im* SCAN
                                                                                               33:20 TIME
                                         Figure  2.  Mass  chromatogram  of m/e 252.

-------
151




2ft? .















1
1
1
1
i
117 II





















r ' •
i!"
!













: t
i «
X
< «i>
. lilC
i ti.
i.?a
1 |jr
i lit
!|
I1!
iifi




1
,1
MI
•1
1 i
i !!
I Si

2:


1
1
1
|
I
16
tian
\
897


l»



[
i "I
I G
•l!ila
'li;!
1 •*
:!
, i
W
!P

\
\
\
|'ji:!M
Ij.Wi
HE
t'lK:
i|»![
! ii-Ulb
il!«'?IH
a
i
«Ll 	 	 ~r— • 	 • 	 , 	 • 	 , 	 	 ; 	 	 f- r-
1
I .^.


019
I! I! 
-------
                                                          738
00
                                                                                   w.ee
                          Figure 4.   Total  ion chromatogram of a typical  coke  oven  sample.

-------
                                                     454
VO
                               164
                                  2*9
                                  6:49
                                                                          T
                                                13:,
                                                                               2f.: 49
                                         Figure 5.  Mass chromatogram of m/e 202

-------
                           IW.t-i



                                I
ro
o
                           50.9-
                           it/T-
                          ier*
                                     57.1
            l<>U


 TV   88.1
^..InJjU.ii..,	h^
     8*      l««
                                              ?81.3
                                                                                        18*
                                                                       34»      36B
                                                                                        389
TT
2;-»
                                             Figure  6.   Mass  spectrum  - scan number 454

-------
ro

                           IVE

                           ISO.*-,
                                              73.1




                                                79.1
                                                 2B»       :y»
                                                                                                        2lt2.3
                                                                          135.2
I"    1R3.2 175.2     .,
,1)111,1,.,.,. j','. j i .-|T|'i'|-. ; n'jTl'l

     160      199      :
                                                                                                               2ie.3
                                                                                                                120
                                                  Figure 7.   Mass spectrum -  scan  number 476.

-------
IN}
INJ
                                   59.
                             «••*
                        IfE
                                     6«       f»
                                                                                                  2H7.2
                                                              120      14*       If*      If*
                                                                                                        223      2*9
                               252.4
                       M/E
                                    268
                                            289
                                                                          isy.6.    3F.3.8    ^83
                                                             329      349
                                                                                       T86
                                              Figure 8.   Mass  spectrum -  scan  number 738.

-------
                                                                                 71(1
ro
OJ
                      252
                                                                                   7TT.

                          _35     .J_M._	f
                                                                                                        . ?• TIIU.
                                            Figure 9.   Mass  chroma to gram  of m/e  252

-------
       252
50.?-
                                                            2M     220
           I   '    I
          26*     2M
                               3»     3«     36*     3M
                 Figure  10.   Mass spectrum - scan number 710.

-------
                                       fcIC
                                       •4/U/79 IS:tl:M
                                       SAHPLE: 3UI BS07C1B
                                       BAMGE: C  1.1122  LABEL
                       DATA:
                       CALI:
BS07CIB It
DC9410 II
SCAMS   I TO 1122
N •. 4.0  QUAN:  A  0.  l.e BASE: U 28.  3
                                                                                     7 7
tn
                          tic
                                            6:40
                                                           13:2*
                                                                                          26:40
                                                                                                          IOM
                                                                                                          33:2«
                                                            SCAN
                                                            TltlE
                                Figure  11.    Total  ion  chromatogram of detection limit test sample.

-------
HASS	
M/ia/79 15:0M
SATtPLE: 30L BS07CIB
RANGEt U  1.1122  LABEL: N 9.
DATA: BS07CIB II
CALI: DCW18 II
                                                                                                          SCANS   I TO 1122
                                                                    WAN: A  •.  1.0  BASE:  U
                       1M
                                                                                      T
ro
                                                                                                                           7832.
                                                                                                                         2*2.«6»
                                                                                                                         *  *.5M
                                                                                                                          SCAN
                                                                                                                          Tint
                                                   Figure  12.   Mass  chromatogram of m/e  Z02.

-------
              HASS CHROnATOGRAH
              W/M/79 15:01.M
              SAHPU: 3UL BS07CIB
              BANGE: G  1.1122  LABEL:
JATA:
CALI:
BS07CIB II
DCMW II
SCANS    I TO 1122
                                    N  •
It*
                                             QUAH: A  •.  l.»  BASE: 0 M.  3
252
                                                                                                    1136*.
                                                                                                   252.W5
                                                                                                  *  «.5M
                   6:4«
                                   13:2*
                                                                     26:40
                                                                                     33:28
                                                                                                    SCAN
                                                                                                    TII1E
                      Figure  13.   Mass chromatogram of m/e  252.

-------
r\>
<»
                           100.
                           300
                                          HAS5 CHBOWTOGIAM                                 DATA: BS07CIB II
                                          04/10/79 I5:0I:0«                                 CAL1: DGMte II
                                          SAIffLE: 3UL BS07CIB
                                          RANGE: G   1.1122  LABEL: N 9. 4.0 QUAN: A  «. !.•  BASE: U 20.   3
                                 SCANS    1 TO 1122



208



131




y
320
3



1

ill
t
728
627
1"
IT in
tj.
1
                                               2M
                                               6:4*
                                                                13:20
20:00
                  8W
                 26:48
                                                                                                                                    340.
                                                                                                                                 300.090
                                                                                                                                *  0.500
                                                                                                                            1096
SCAN
TItlE
                                                     Figure 14.   Mass chromatogram  of  m/e  300.

-------
HASS CHNHATOGRAn
M/10/79 15:«1:M
SAHHE: 3UL BS07CIB
BANGE: G  1.1122  LABEL: N
                                                                ». 4.«  (MAN
                                                                                      DATA:  BS07C1B II
                                                                                      CALI:  DC0410 II
                                                                                   •  BASE: U 28.  3
SCANS    1 TO 1122
ro
                                                                                                                              351.
                                                                                                                             SCAN
                                                                                                                             TINE
                                                  Figure  15.   Mass  chromatogram of m/e  302.

-------
            MASS GHROttTOGBAH                               DATA: BS07CIB II
            M/I9/79 I5:«1:M                               CALI: BCMie II
            SAHTLE: 3UL BS07CIB
            RANGE. G  1.1122 LABEL: N  1,,4.9 QUAH: A  1.  1.9  BASE. 0 ».  3

                                     7584.
                                     ~ 912.
                                                     467
SCANS  4M TO  5»
                                                                                                   7584.
                                                                                                 2t2.«6«
                                                                                                * «.5M
13:2*
                                                                              16:49
                                                                                              52»  SCAH
                                                                                             17:20 TINE
                        Figure 16.   Mass  chromatogram  of  m/e 202.

-------
HASS CHBOMATOGRAh
04/10/79 15:01:00
SAMPLE: 3UL BS07CIB
BANCE: G 68*. 740 LABEL:
DATA:
CALI:
                                                                   BS07CIB 1467
                                                                   BC04I0 II
SCANS  650 TO  800
                                     N  1. 4.8 QUAN:
                                                         1.0  BASE: U 20.  3
100.
252
                                                                                                       6784.
                                                                                                     252.075
                                                                                                       0.500
                                                                                                       SCAN
                                                                                                 26:40 Tlffi
                             Figure  17.    Mass  chromatogram  of m/e  252.

-------
                    5.0  SUGGESTIONS  FOR FUTURE  STUDY
     Other alternatives exist to the  screening technique proposed  in the
Arthur D. Little, Inc.  report.   High  performance liquid chromatography
(HPLC) is a technique that can  provide  a semi-qualitative and quantitative
analysis for PAH in a time span as  short as  15 minutes.  As with the
gas chromatograph, the  identity of  the  eluting components are assigned by
retention time and spiking.   Specific detection  techniques  such as
ultraviolet absorption  and fluorescence can  be employed to  screen  for PAHs.
The selectivity of these detectors  are  extremely useful when analyzing
complex samples.  Setting the UV or fluorescence detector at wavelengths
specific for the PAH(s) provides for an enhanced response of the components
of interest over other components in the sample  matrix.
     Figures 18 through 20 are high performance  liquid chromatograms of
PAH standards and a coke oven sample.  Figure 18 is a  standard mixture
of benzene, chloronaphthalene and benzo(a)pyrene.  Under  these conditions,
20% water 1n methanol at a flow of 1  milliter per minute  on a  25 cm
ODS column, benzo(a)pyrene elutes 1n approximately 35  minutes  while
benzene elutes in 7 minutes.  Figure 19 1s the chromatogram of naphthalene,
fluorene, phenanthrene and benzo(a)pyrene.  In this case  1t takes  70
minutes for benzo(a)pyrene to elute from the column.  This  longer  analysis
time allows for better Identification of components in complex mixtures.
Figure 20 1s a coke oven sample run under the same conditions  as  the
standard 1n Figure 19.  As can be seen, this provides  an  acceptable
separation of a very complex sample.  The presence of fluorene and
phenanthrene can  be easily distinguished.
     These examples are typical of the results that can be obtained using
HPLC and a UV spectophotometer detector.
      Unlike nonaqueous reverse  phase chromatography (NARP), conventional
reverse  phase high performance  liquid  chromatography  involves the  use of
nonpolar chromatographic  stationary phase (e.g.  c-18  bonded to an  inert
support) and a  polar solvent system  (e.g. methanol/water).  In the case of
PAH compounds,  the use of this  type of  system for  separation of PAHs above
300 molecular weight can  result  in excessive analysis time.  This  is due
in part  to the  limited solubility  of PAHs in the solvent systems  used.
Substituting a  less  polar solvent  system  for the polar solvent systems

                                     32

-------
                    5.0  SUGGESTIONS  FOR FUTURE  STUDY
     Other alternatives exist to  the  screening technique proposed in the
Arthur D. Little, Inc.  report.   High  performance liquid chromatography
(HPLC) is a technique that can provide a semi-qualitative and quantitative
analysis for PAH in a time span as  short as  15 minutes.  As with the
gas chromatograph, the  identity of  the eluting components are assigned by
retention time and spiking.  Specific detection  techniques  such as
ultraviolet absorption  and fluorescence can  be employed to  screen for PAHs.
The selectivity of these detectors  are extremely useful when analyzing
complex samples.  Setting the UV or fluorescence detector at wavelengths
specific for the PAH(s) provides for an enhanced response of the components
of interest over other components in the sample  matrix.
     Figures 18 through 20 are high performance  liquid chromatograms of
PAH standards and a coke oven sample.  Figure 18 1s a  standard mixture
of benzene, chloronaphthalene and benzo(a)pyrene.  Under  these conditions,
20% water in methanol at a flow of 1  milliter per minute  on a  25  cm
CDS column, benzo(a)pyrene elutes 1n approximately 35  minutes  while
benzene elutes in 7 minutes.  Figure 19 1s the chromatogram of naphthalene,
fluorene, phenanthrene and benzo(a)pyrene.  In this case  1t takes  70
minutes for benzo(a)pyrene to elute from the column.  This  longer analysis
time allows for better identification of components in complex mixtures.
Figure  20 1s a coke oven sample run under the same conditions  as  the
standard 1n Figure 19.  As can be seen, this provides an  acceptable
separation of a very complex sample.  The presence of fluorene and
phenanthrene can  be easily distinguished.
     These examples are typical of the results  that can be obtained using
HPLC and a UV spectophotometer detector.
      Unlike nonaqueous reverse phase chromatography (NARP), conventional
reverse phase  high performance  liquid chromatography  involves the  use of
nonpolar chromatographic  stationary  phase (e.g.  c-18  bonded to an  inert
support)  and a  polar solvent system  (e.g. methanol/water).  In the case of
PAH  compounds,  the use of this  type  of system for  separation of PAHs above
300  molecular weight can  result  in excessive analysis time.  This  is due
in part to  the  limited solubility  of PAHs in the solvent systems used.
Substituting a  less  polar solvent  system  for the polar solvent systems

                                     32

-------
routinely used in reverse  phase  chromatography  Increases the solubility of
the PAHs.  Consequently, with  the  use of methylene chloride and acetonitrile
as a solvent system, coronene  (m/w 300) and  decacyclene  (m/w 454) can be
eluted in approximately 8  and  17 minutes,  respectively.  Thus with NARP,
components which cannot be analyzed by  GC/MS can  be  handled easily and
in a relatively short analysis time.  It is  recommended  that HPLC be
investigated as an additional  method for the analysis  of PAHs.
                                     33

-------
        START
CO
                                 figure 18.  High performance liquid chromatogram
                              of A. benzene, B. chlorobenzene and C. benzo(a)pyrene.

-------
Ut
cn
                      . I

                       4
                    ___ |__
                     ] __3..
                           -O-
                                                         -5O-
                                 Figure 19.  HPLC of A. naphthalene,  B.  fluorene,
                                      C. phenanthrene and D.  benzo(a)pyrene.

-------
CO
CT>
                                        Figure 20.   HPLC of coke oven sample
                                              A.  fluorene; B. phenanthrene.

-------
                              6.0   CONCLUSION

     It is clear from the data presented  1n  this  report that the spot test
has proven acceptable as  a screening  tool  for  samples  submitted for GC/MS
analysis of PAHs.  Two requirements were  to  be met  by  this  screening
procedure.  One, the detection limit  of the  PAH spot test must be  below
that of the GC/MS and two, the interferences that prevent accurate results
from the spot test must be defined.  These issues have been addressed.
That is, the detection limit of the screening  technique  (spot test) is
below that of the accepted analysis technique  (GC/MS). Also, interferences
occuring from the inherent color of the sample were discussed.
     It was shown that unconfirmed positives can  be obtained  from  several
sources.  A unconfirmed positive result will occur  when  a  sample contains
a level of PAHs which is  above the detection limit  of  the  spot  test and
below the detection limit of the GC/MS.  Another  source  of unconfirmed
positives can be caused by the interference of the  sample  color with  the
fluorescence of the sample.
     Another source of false positives which must not  be overlooked  are
cases in which the fluorescence is caused by PAHs which  have a  molecular
weight above 300.  GC/MS will not prove adequate  as a  method of analysis.
Clearly, high performance liquid chromatography must be  used in cases where
high molecular weight PAHs are suspected.   Indeed,  it  may  prove beneficial
to analyze all samples which prove positive by the spot test by HPLC for
high molecular weight PAHs and not only those  which prove  positive by the
spot test and negative by GC/MS.
                                     37

-------
                         7.0  REFERENCE
1.  Sensitized Fluorescence for the Detection  of Polycyclic
    Aromatic Hydrocarbons,  Report No.  EPA-600/7-78-182,  NTIS
    No. PB 287-181  IAS, September 1978.
                                38

-------
                               TECHNICAL REPORT DATA
                         (Please read Inunctions on the reverse before completing)
  REPORT NO.
 EPA-600/7-79-207
                          2.
                                                     3. RECIPIENT'S ACCESSION NO.
 TITLE AND SUBTITLE
Evaluation of Sensitized Fluorescence for Polynuclear
 Aromatic Hydrocarbon Detection
                                                      REPORT DATE
                                                     August 1979
                                                    6. PERFORMING ORGANIZATION CODE
 . AUTHOR(S)

T.R. Smith
                                                     8. PERFORMING ORGANIZATION REPORT NO.
.. PERFORMING ORGANIZATION NAME AND ADDRESS
TRW Defence and Space Systems Group
One Space Park
Redondo Beach, California 90278
                                                     10. PROGRAM ELEMENT NO.
                                                     INE624
                                                     11. CONTRACT/GRANT NO.

                                                     68-02-2689, T.D.  104
 12. SPONSORING AGENCY NAME AND ADDRESS
 EPA, Office of Research and Development
 Industrial Environmental Research Laboratory
 Research Triangle Park, NC  27711
                                                     13. TYPE OF REPORT AND PE
                                                     Task Final; 1-6/79
                                                                      PERIOD COVERED
                                                     14. SPONSORING AGENCY CODE
                                                      EPA/600/13
5. SUPPLEMENTARY NOTES
 919/541-2557.
                              project officer  is Larry D. Johnson, Mail Drop 62,
16. ABSTRACT rphe repOrt gives results of an evaluation of a fluorescent spot test for
detecting the presence of polynuclear aromatic hydrocarbons  (PAHs) as a screening
technique for samples to be analyzed by Gas Chromatography/Mass Spectrometry
(GC/MS).  The test is based on the phenomenon of sensitized fluorescence and is
capable of easily detecting 100 picograms  of PAH in a 1 microliter sample, a level
of sensitivity adequate for screening combustion effluent samples. Two interferences
were observed: highly colored samples require  dilution to allow viewing of the fluor-
escence level, and samples containing substantial amounts of phthalate esters pro-
duce false positive results. No false negative  results were observed in the study.
The test is adequate for screening combustion effluent samples prior to GC/MS
analysis.
 7.
                             KEY WORDS AND DOCUMENT ANALYSIS
                 DESCRIPTORS
                                          b. IDENTIFIERS/OPEN ENDED TERMS
                                                                    COS AT I Field/Group
 Pollution
 Combustion Products
 Screening
 Aromatic  Poly eye lie Hydrocarbons
 Fluorescence
 Sensitizing
                                         Pollution Control
                                         Stationary Sources
                                         Sensitized Fluorescence
13B
21B
14B
07C
20F
13H
"S. DISTRIBUTION STATEMEN1

  Release to Public
                                         19. SECURITY CLASS (ThisReport)
                                          Unclassified
21. NO. OF PAGES
       43
                                         20. SECURITY CLASS (This page)
                                          Unclassified
22. PRICE
 EPA Form 2220-1 (9-73)
                                         39

-------