NBSIR 76-1090
EPA 550/9-76-009
Environmental Effects on
Microphones of Various
Constructions
Gale R. Hruska, Edward B. Magrab, William 8. Penzes
Institute for Basic Standards
National Bureau of Standards
Washington, D. C. 20234
July 1976

Final Report
Prepared for
U. S. Environmental Protection Agency
Office of Noise Abatement and Control
Washington, D. C. 20460

-------
NBSIR 76-1090
EPA 550/9-76-009

ENVIRONMENTAL EFFECTS ON
MICROPHONES OF VARIOUS
CONSTRUCTIONS
Gale R. Hruska, Edward B. Magrab, William B. Penzes
Institute for Basic Standards
National Bureau of Standards
Washington. D. C. 20234
July 1976

Final Report
Prepared for
U. S. Environmental Protection Agency
Office of Noise Abatement and Control
Washington, D. C. 20460
  •t--. _ -* •*
  •*•,»„••
 U.S. DEPARTMENT OF COMMERCE. Elliot L Richardson. Secretary
  Dr. Betay Ancfcer-Johnwn, AutsUutt Secretary tor Science end Technology
 NATIONAL BUREAU OF STANDARDS. Ernest Ambler, Acting Director

-------
                                ABSTRACT





     The pressure sensitivities of two "1/2-inch" electret,  two "1-inch"




ceramic, and two back-vented "1-inch" condenser microphones  were measured




for numerous combinations of temperature, percentage relative humidity,




and frequency.  The two condenser microphones were calibrated by the reci-




procity technique at each combination of temperature, relative humidity




and frequency.  The condenser microphones were then used as  calibrated




sources to determine the pressure sensitivities of the other microphones.




Insert voltage techniques were used to eliminate the environmental effects




on the electronics.  It was found that the back-vented condenser micro-




phones are insensitive to changes in relative humidity.  At  frequencies




considerably below their resonance frequencies they exhibited only a very




small change in sensitivity with temperature.  At frequencies closer to




the resonance frequency the temperature coefficient increases approximately




fourfold.  The temperature and humidity coefficient for the  electret and




ceramic microphones could not be determined due to the instability in their




sensitivities which produced changes that were larger than those induced




by the temperatures and humidities.
Key Words:  Calibration; ceramic; condenser; electret; humidity;




            microphones; reciprocity; sensitivity; temperature.

-------
                              INTRODUCTION


     The introduction of federal regulations stipulating the permissible

noise levels in the environment has made it necessary for many acoustic

measurements to be performed over extended periods of time during which

the temperature and humidity vary markedly.  These changes in the environ-

mental conditions could affect the sensitivity of the microphones.  There

exist some published data1"5 concerning the effects of environmental con-

ditions on microphones of various construction.  However these data appear

in abbreviated form, usually in terms of frequency-independent temperature

coefficients.  Furthermore the experimental techniques employed to obtain

this information are rarely stipulated.  It was felt therefore that a

modestly thorough investigation, which determined the changes in the pres-

sure sensitivity of those commercially available microphone constructions

most frequently found in noise-measuring and noise-monitoring systems and

which used a consistent and standardized measurement procedure would pro-

vide useful and meaningfully-comparable data.

     Six commercially-available microphones were measured:  two "1/2-inch"

electret, two "1-inch" ceramic, and two "1-inch" condenser microphones.

The condenser microphones had their pressure equalization port back-vented

through a dehumidifier containing silica gel.*  The sensitivity was meas-

ured (where physically possible) at the following combinations of frequency,
 *By  their very nature the small volumes in the electret and ceramic micro-
 phone  cartirdges are unaffected by humidity in the same way that condenser
 microphones are and therefore do not require a desiccant.  However, certain
 electret or ceramic configurations protect their electronics from the
 effects of humidity with a desiccant.

-------
temperature, and relative humidity:  frequency: 0.1, 0.2, 0.5, 1, 2, 3, 4,




and 5 kHz; temperature: every 10 °C from -20 to +50 °C; and  (nominal) rela-




tive humidity: every 10% RH from 25 to 95% RH.





                        TEST METHOD AND PROCEDURE




     The sensitivities of the microphones were measured using the procedures



specified in the U.S. Standard.6  The NBS facility used to calibrate micro-




phones was essentially duplicated, except that the 3-cm3 coupler, the




microphones and the microphone preamplifiers were placed in an environmental




chamber having wet and dry bulb temperature control and spatial uniformity




to within ± 0.2°C.   So that the microphone could be inserted into the




coupler, several different size adapter rings, which are used to electri-




cally isolate one of the microphones in the coupler, were required to com-




pensate for thermal expansion and contraction over the temperature range.




The microphone preamplifiers, having insert voltage capabilities, provided




a means whereby the effects of the environmental conditions on the pre-




amplifiers were eliminated.




     The coupler, the preamplifiers and the six microphones (plus an




additional "1-inch"  condenser microphone required  for  the reciprocity




calibration) were placed in the environmental chamber.  The  desired  tem-




perature and relative humidity were set and then two hours were  allowed




for the chamber  to reach equilibrium.  Then the "source" microphone  was




placed in the  coupler and the "receiver" microphone connected to the pre-




amplifier but  not placed in the top of the coupler.  An additional fifteen




minutes were allowed for the chamber to return to  equilibrium at which time




the receiver microphone was rapidly placed into the coupler.  After

-------
equilibrium had again been established, the measurements were started by




first determining the resonance frequency of the lowest longitudinal mode




of the coupler volume.  Then the voltage ratio measurements  were made in




the following order:  5, 0.1, 0.2,...4, and 5 kHz.  Then the resonance fre-




quency measurement was repeated.  By examining the first and final measure-




ments at both 5 kHz and the two resonance frequencies, it was possible to




check that equilibrium conditions existed during the course of the




measurements.




     The sequence of measurements was as follows:  For a given temperature




and humidity the three condenser microphones, labeled A, B, and C, were




interchanged after each frequency run,as shown in the table below:




                  Frequency Run     Source     Receiver




                        1             A           B




                        2             AC




                        3             B           C







Runs  1 and 2 gave the ratio of  the  responses of  the microphones B  and C.




Using the results of Run 3 plus  the measurement  of the capacitance8  of




microphone B,  the absolute pressure response levels of microphones A, B,




and C were calculated.  The pressure response of  the remaining  four micro-




phones was determined using microphone B as the  source, with the measure-




ments on the electret microphones preceding those on  the  ceramic micro-




phones.  More  specific  details  of the procedure  are given  in Appendix A




along with a discussion  of the  uncertainties in  the measurement.   The




total time, after warm-up, to perform the  reciprocity and  comparison meas-




urements on all  six microphones at  a  given temperature and humidity  was




approximately  6  hours.   The  temperatures were  changed in  the following

-------
order:  20, 30, 40, 10, 0, -10, -20, 50, and 20°C.  At each temperature




the humidity was Incrementally increased from the lowest to the highest




obtainable humidity.  The lowest humidity depended on the temperature,




with this minimum value increasing with decreasing temperature.  Below




0°C the relative humidity could not be determined with good precision and




only one set of microphone calibrations was made at these temperatures.




The repeated calibration at 20°C was only performed at 44% RH.





                          DISCUSSION OF RESULTS




     The changes in sensitivity of the six microphones normalized to




their respective pressure sensitivity at 1 kHz and 20°C are shown in




Figs. 1 through 8.  Figures 1 and 3 show the change in sensitivity of




the "1-inch" condenser microphones as a function of frequency for the




various temperatures.  These figures are replotted in Figs. 2 and 4,




respectively, as a function of temperature for selected frequencies.




The data points shown in these four figures are the mean values of the




sensitivity changes for the range of humidities tested.  The spread of




the data at virtually every one of the points is less than ± 0.15 dB.




These figures show that the two "1-inch" condenser microphones exhibited




small temperature coefficients (change in microphone sensitivity per




change in temperature) of approximately -0.003 and +0.005 dB/°C, respec-




tively, from -20 to 50°C and for frequencies below 1000 Hz.  At frequen-




cies above 1000 Hz the temperature coefficient increased, such that at




5000 Hz they become approximately -0.02 and -0.03 dB/°C, respectively.




This variation in the temperature coefficient as a function of frequency




is due to the relatively large amount of viscous damping introduced by

-------
the particular backplate design of these microphones, that is, by the




distance between the diaphragm and the backplate and the location, number,




and size of the holes in the backplate.  Since the viscous losses greatly




affect the sensitivity near resonance, smaller microphones could be ex-




pected to show less sensitivity change at high frequencies.




     The changes in sensitivity for the electret microphones are shown in




Figs. 5 and 6 and for the ceramic microphones in Figs. 7 and 8.  In these




figures the vertical bars indicate the range of the data as a function of




humidity.  The absence of the vertical bars indicates that the data were




only taken at a single relative humidity.  The data were presented in




this form because no meaningful relation could be established for the




change in sensitivity of these microphones as a function of either fre-




quency, temperature or humidity.  This type of behavior can be explained




if the changes due to temperature and humidity are much smaller than the




inherent instability of the microphone itself.  Consequently it is not




possible to obtain from these measurements the temperature coefficients




of the electret and ceramic microphones.



      If  the electret and ceramic microphones  are subject  to these in-




stabilities they will exhibit  similar  behavior at  standard  conditions,




20  °C and  44% RH.  As can be seen  from Tkbles 1-3  the  condenser micro-




phones are within  ±0.2  dB of their original  sensitivities whereas the




electret and  ceramic microphones  show  relatively  large variations.   These




tables give  an indication of the  moderately  long-term stability  of  these




types of microphone  constructions.

-------
                               CONCLUSIONS




     From the data presented the following conclusions are reached:




     1.   The back-vented condenser microphones with a dehumidifier were




insensitive to relative humidity.




     2.   At frequencies far below the condenser microphone's resonance




frequency the temperature coefficient was extremely small (^ ± 0.005 dB/°C),




     3.   Condenser microphones containing a large amount of viscous




damping would be expected to have a greatly increased (by a factor of A)




temperature coefficient in the vicinity of the microphone's resonance




frequency.




     4.   The electret microphones which were tested exhibited short-term




sensitivity instabilities of the order of ± 0.5 dB.  The ceramic micro-




phones examined showed short-term instabilities of the order of ± 1 dB




or larger, at some frequencies.  The magnitudes of these instabilities




made it impossible to determine the changes in sensitivity as functions




of temperature or relative humidity.




     5.   The condenser microphones exhibited long-term instabilities




(i.e., over a period of about sixteen weeks) of the order of ± 0.2 dB.




The electret and ceramic microphones showed long-term instabilities of




up to ±1.5 dB, with larger values at a few frequencies.





     From the data which were taken, it is not possible to predict with




assurance what uncertainties would occur if the ceramic and electret




microphones were used for outdoor noise measurements.  It seems reason-




able, however, to assume that if significant changes in temperature and




relative humidity occur between system calibrations, then the uncer-




tainties in the measured sound levels could be at least of the magnitude

-------
of the short-term instabilities already mentioned.   Furthermore,  in actual




sound measuring systems, changes in temperature and relative humidity




could affect the electronics thus introducing additional changes  in over-




all system sensitivity.

-------
                               REFERENCES






 1G. M.  Sessler  and J. E. West, "Electret Transducers: A Review," J.




   Acoust.  Soc.  Amer. 5_3, 1589-1600  (1973).




 2L. L.  Beranek, Ed., Noise and Vibration Control  (McGraw-Hill Book Co.,




   New York, 1971) pp. 48-49.




 3G. Rassmussen, "Reliability of Measurement Microphones Under Outdoor




   Conditions,"  Seventh International Congress on Acoustics, Budapest, 1971,




   pp. 289-292.




 ^Rassmussen, "Measurement Microphones," Inter-Noise 74 Proceedings,




   International Conference on Noise Control Engineering, Washington, D.C.




   (September 1974), pp. 55-60.




 5A.P.G. Peterson and E. E. Gross, Jr., Handbook of Noise Measurement, 7th




   Edition  (General Radio Co., Concord, Mass., 1972), pp. 172-173.




 6"Method for the Calibration of Microphones," SI.10-1966, American National




   Standards Institute, New York, N. Y.




 7R. K.  Cook, "Measurement of Electromotive Force of a Microphone," J. Acoust.




   Soc.  Amer. 19_, 503 (1947).




 8W. Koidan, "Method for Measurement of |E'/I'| in the Reciprocity Calibration




   of Microphones," J. Acoust. Soc. Amer.  32, 611 (1960).
USCOMM-NBS-DC

-------
                       TABLE 1

Long-Term Stability of "1-inch" Condenser Microphones
                 at 20°C and 44% RH
                         Change in Sensitivity
                     (dB re value on May 21, 1975)
lency (Hz)

100
200
500
1000
2000
3000
4000
5000

100
200
500
1000
2000
3000
4000
5000
June 17,

0.17
0.07
0.04
0.06
0.07
0.08
0.08
0.07

+0.18
+0.17
+0.18
+0.12
+0.01
-0.09
-0.15
-0.17
1975 July 22, 1975
Cartridge //I
-0.11
-0.12
-0.15
-0.12
-0.03
+0.11
+0.17
+0.21
Cartridge #2
+0.18
+0.12
+0.06
+0.06
+0.03
+0.05
+0.02
+0.03
September 9, 1975

-0.04
-0.09
-0.09
-0.05
+0.01
+0.11
+0.19
+0.22

-0.11
-0.16
-0.15
-0.12
+0.01
+0.13
+0.18
+0.26

-------
                        TABLE 2


Long-term Stability of "1/2-inch" Electret Microphones

                  at 20°C and 44% RH
                          Change in Sensitivity
                      (dB re value on May 21, 1975)
luency (Hz)

100
200
500
1000
2000
3000
4000
5000

100
200
500
1000
2000
3000
4000
5000
July 23, 1975
Cartridge //I
-1.22
-0.51
-0.30
-0.34
-0.34
-0.38
-0.40
-0.32
Cartridge #2
+0.26
-0.02
-0.11
-0.12
-0.12
-0.14
-0.16
-0.17
September 9, .

+1.53
+1.33
+1.25
+1.24
+1.17
+1.05
+0.97
+0.93

+2.47
+1.26
+0.88
+0.88
+0.88
+0.84
+0.70
+0.59
                           10

-------
                      TABLE 3

Long-Term Stability of "1-inch" Ceramic Microphones

                at 20°C and 44% RH
                        Change in Sensitivity
                    (dB re value on May 21, 1975)
luency (Hz)

100
200
500
1000
2000
3000
4000
5000

100
200
500
1000
2000
3000
4000
5000
July 23, 1975 September 9, :
Cartridge #1
-1.51
-1.47
-1.44
-1.45
-1.40
-1.40
-1.56
-3.90
Cartridge #2
0.23
0.24
0.26
0.30
0.49
0.78
0.52
0.07

0.19
0.04
0.04
0.04
0.31
0.81
1.16
0.70

0.73
0.12
0.00
0.05
0.57
1.08
1.25
0.36
                         11

-------
o
o
O
CM
|i -'
UJ i_
co <
UJ >- -2
Z 1-
o ^:
•^ ^
MICROP*
SENSITI
*
0) • A
^T
GQ
-5
0.
1
3 § 9
0 6 o V
o o
a _,
o
9
o -2
o IO°C A
o -3
A 0°C
• v-|Q°C
a-20°C -4
-i 	 J 	 J i 	 i ill 5

fi 8 6
-v A
®
i
a
o 50°C
A 40°C D
- v 30°C X
a 20°C
i i i i i i i 1 1 1 i iii
1 0.2 0.5 1 2 5 O.I 0.2 0.5 1 2 5
       FREQUENCY (kHz)
FREQUENCY (kHz)
FIGURE 1,  CHANGE IN SENSITIVITY OF  "I-INCH" CONDENSER MICROPHONE NO, 1
          AS A FUNCTION  OF FREQUENCY AND TEMPERATURE,  THE CHANGE IN
          RELATIVE HUMIDITY INTRODUCES AN UNCERTAINTY ABOUT THE DATA
          POINTS OF LESS THAN * 0,15 DB,

-------
u>
1
o
o
^ 0
u
tS
^ <
^
ul Is1
t x -i
CO -*
z —
UJ i_
CO <
UJ > ~2
Z 1-
i>
0- h-
g»-3
O S
— UJ
2 co

-------
1
o
e
0
>Q °
li -'
tu Z
CO 5
UJ >- -2
o 5
O- P
— UJ
2 CO
O> —4
CD
5
1
'- ? 8 I o
0 ""'
V
A
o
D 	 £
o |Q°C v
A 0°C o
. v -|Q°C ° ,
a -20°C v *
A
o
-4
J 	 1 	 L 11 	 | ill e
O O
"A y A
•' ' • 1
1
b
J
o 50° C
A 40° C
v 30° C
D 20°C I
0

a
V
A
r
O
1 	 i i i i i t i i i i , , >
O.I 0.2 0.5 1 2 5 0.1 0.2 0.5 1 2 5
      FREQUENCY  (kHz)
FREQUENCY  (kHz)
FIGURE 3,  CHANGE  IN SENSITIVITY OF "I-INCH" CONDENSER MICROPHONE NO, 2
          AS A FUNCTION OF  FREQUENCY AND TEMPERATURE,  THE CHANGE IN
          RELATIVE HUMIDITY INTRODUCES  AN UNCERTAINTY ABOUT THE DATA
          POINTS  OF LESS THAN ± 0,15 OB,

-------
                                         200 Hz
   -20
FIGURE
  -10
0        10      20       30

 TEMPERATURE  (°C)
CHANGE IN SENSITIVITY OF "I-INCH" CONDENSER MICROPHONE NO, 2
AS A FUNCTION OF TEMPERATURE FOR SEVERAL FREQUENCIES,

-------
1
t -\ r\
SENSITIVITY
AT 1 kHz AND 20 °(
iii'
•fr OJ ro — c
ROPHONE
JSITIVITY
3 - rv> rt

-------
^ 2
O
0 1
O
: SENSITIVITY
' AT 1 kHz AND
i i
to — C.
MICROPHONE
i SENSITIVITY
5 — ro
V \J
flQ -1
•o
^ -2
20° C I
T TTiTiiT I 1 n
J[ A | I | |1 | O
-1
-2
FREQUENCY (kHz)
1 1 i
1 1 i
).l 0.2 0.5 1 2 51
30°C T
' i i, -i i1'1 :
i i M ' .,
-2
40° C
1 1 1 T IT Tli
j i 1 1 ij jii
FREQUENCY (kHz)
1 1 1 1 1 II 1 II
1 1 1 1 1 II 1 II
D.I 0.2 0.5 1 2 E
50°C
i i ,,i .i i !l
1 1 i 1 '1 J-
FIGURE SB,   CHANGE IN SENSITIVITY OF "1/2-iNCH" ELECTRET MICROPHONE NO,  1
            AS A FUNCTION OF FREQUENCY AND TEMPERATURE.   THE VERTICAL BARS
            INDICATE THE RANGE OF THE DATA AS A FUNCTION OF  HUMIDITY,

-------
CD
1
O
0 -|
SENSITIVITY
\T 1 kHz AND 20
•*> w ro
CROPHONE !
iNSITIVITY >
O — ro
^^ CO
2 "'
CD -2
^ -3
-4
-20°C ° 2
-1 	 1 1 1 1 I 1 ' oil
1 1 1 1 1 1 1 1 Til
0
0
\J
o
-1
-2
0
-3
FREQUENCY (kHz)
I 	 1 — i — i i i i 1 1 1 	 i i i
1 i i 1 1 i p'i
3.1 0.2 0.5 1 2 5 C
2
-IO°C
0 ,
o
H 	 1 	 1 — 1 1 1 1 1 1 1 	 L ? 1 1 n
i i o111© ofllU
-1
-2
o
o°c
I
1 1 1 1 1 1 1 1 1 ill
1 1 Jill I J III
FREQUENCY (kHz)
i 1 i i 1 1 i iii
1 1 1 1 1 1 1 M 	 1 — | — H
11 0.2 0.5 1 2 5
IO°C T
T i
}i i T i i T T III
" 1 1 ii } [iii
            FIGURE GA,
CHANGE IN SENSITIVITY OF "1/2-iNCH" ELECTRET MICROPHONE NO, 2
AS A FUNCTION OF FREQUENCY AND TEMPERATURE,   THE VERTICAL BARS
INDICATE THE RANGE OF THE DATA AS A FUNCTION OF HUMIDITY,

-------
VO
2
0
0 1
O
n
• SENSITIVITY
r AT 1 kHz AND
i i
ro — <
MICROPHONE
) SENSITIVITY
D — ro
us —
CD -1
^ -2
20°C T 2
I '
T — T 1 1 I 1 1 1 ll I 1 1 O
1 > i • 1 1 I IA | | j u
-\
-2
FREQUENCY (kHz)
1 	 1 	 1 — 1 — 1 1 MM — 1 III
i i 1 1 M 1 1 | III
>•! 0.2 0.5 1 2 51
30°C T 2
	 1 — 1 — Mill 	 1 * 1 n
J^ i i x i i i n 1 1 1 o
-1
n
40°C
i T 1 I T i ' ' T T lit
[ 1 1 1 1 1 1 1 l — T — H
FREQUENCY (kHz)
i i i i i i 1 1 ii
1 1 t 1 Mill 	 1 	 1 — 1 — |
3. 0.2 as 1 2 5
50°C T
i iiliiil .. .11
i t 1 t M 1 11 	 1 — I — H
             FIGURE  SB,   CHANGE  IN  SENSITIVITY  OF  "1/2-iNCH"  ELECTRET MICROPHONE  NO,  2
                         AS A  FUNCTION OF  FREQUENCY AND  TEMPERATURE,   THE  VERTICAL  BARS
                         INDICATE THE RANGE OF  THE DATA  AS A  FUNCTION OF HUMIDITY,

-------







t
CO
UJ
CO
UJ
z
o
CL
O
cr
0
5




3

^ 2
0
• ,
0
CM
0 °
r 3
-20°C
2

i
o o '

1 1 1 1 1 1 1 1 1 1 1 III D
1 1 1 1 1 1 1 1 II 1 III U
O 0 0
-
o°c 1





1 1 1 1 1 I 1 NT



L















> ^ 1 1 j. 1 1 1 11 -•-
1
^^
^
<
>-
1 1 1 1 1 1 1 1 1 1 1 1 1 J 1 1 1 1 1 1 1 1 1 1
1 1 1 1 1 1 1 1 1 1 1 1 1 ] \ 1 1 1 1 1 1 H



1- O.I 0.2 0.5 1 2 5 O.I 0.2 0.5 1 2 J
> FREQUENCY (kHz) FREQUENCY (kHz)
l_
co 3
QJ 2
CO

-------
3
^ 2

0 '
CM
>0 °
—
20 °C

-
-
T ' 1 1 -r 1 1 1 1
T j i ijiinj
^ < -|L
P £
CO •*
UJ |_
CO <
I I 1 1 1 1 1 1 1
1 1 1 1 1 1 1 1 1
UJ >- O.I 0.2 0.5





ii

]
-

L







[




















2

]
M









[
•

2

1
0

5 -1
gt FREQUENCY (kHz)
^h^^ ^^^fc
Q- H
055 3r
z
i co 2
. 1
§ °
-1
30 °C
—
-
Ii i i i i i 1 1
i ' 'i""
-





•







•
•







» •









«


«
-2

-3

40 °C

^m
1 1 1 1 1 1 T



•^
-L
-
-L
J-

~4"

1 1 1 1 1 1
1 1 till
O.I 0.2 0,5 1 2 I
FREQUENCY (kHz)

FIGURE ZB,   CHANGE IN SENSITIVITY OF "I-INCH" CERAMIC MICROPHONE NO,  1
            AS A FUNCTION OF FREQUENCY AND TEMPERATURE,   THE VERTICAL BARS
            INDICATE THE RANGE OF THE DATA AS A FUNCTION OF HUMIDITY,

-------
to
NJ
0
o
O
CVJ
> o
> <
A B^^l
CO *
UJ i_
co <
UJ >-
1>
0- h-
21
— LJ
2 co


-------
to
1
O °
o
O -i
SENSITIVITY
M 1 kHz AND 2
i i i
•fr w ro
MICROPHONE !
SENSITIVITY )
o -
m~ -~*
9
»• -1
ffi
S -2
-3
- -20°C 0 2
1 1 1 1 1 1 1 1 1 1 ? 1 ? 1
1 1 1 1 1 1 1 1 II 1 111
0 6
0 o
O u
-1
-2
0
FREQUENCY (kHz)
1 	 1 	 1 	 i 1 1 1 1 1 1 1 III
i i i i i i i i i i 1 111
).l 0.2 0.5 1 2 5 (
2
- HO°C ,
-| 	 1 — i i i i i i 1 1 i ill n
1 1 1 1 1 1 1 1 1 1 1 Oil U
00°
0 -i
o
-2
0
-3
o°c
I l 1 I i l 1 1 IT -l
•
FREQUENCY (kHz)
Ii i i i i 1 1 1 1 i
i i l l l 1 1 1 1 	 1 — l — l — l
D.I 0.2 0.5 1 2 I
io°c T
-1 1 1 l l i i i iT
i ±''±""1 -LJ-
           FIGURE  SA,
CHANGE IN SENSITIVITY OF "!-INCH" CERAMIC MICROPHONE NO, 2
AS A FUNCTION OF FREQUENCY AND TEMPERATURE,   THE VERTICAL BARS
INDICATE THE RANGE OF THE DATA AS A FUNCTION OF HUMIDITY,

-------
to
o 2
o
O
CVJ
> 0 o
£<
t£
20°C
_
T -r 1 1 T 1 1 1 IT
T 1 1 J 1 1 1 IJ.

CO -* _2-
LU )_
CO <


I I







I,
j. \
i




uj > FREQUENCY (kHz)
2 I-
i i i i i 1 1 1 1
i i i i i 1 1 1 1
O > O.I 0.2 0.5 1
0- P
21 •
2g
2 co o
4)
fc- _|
CD
•O -2
30°C
IT 1 1 1 1 1 1 l-r
T T


1
1
2

I'T"




I
I
5


T i
1
ll


2
1
0
-I

-2

-3
40°C
_
i i iiiiii-i- T T'i
1 1 1 J- 1 1 1 IT _L J. -r 1

-

-
FREQUENCY (kHz)

1 1 1 1 1 1 1 1 1 1 III
1 1 1 1 1 1 1 1 r 1
0,1 0.2 0.5 1 25
2
1
0
1

-2
^™
50°C
. i i i i i 1 1 1 i iii
I x J I Iji

-
             FIGURE SB.   CHANGE IN SENSITIVITY OF  "I-INCH"  CERAMIC  MICROPHONE  NO,  2
                         AS A FUNCTION  OF  FREQUENCY  AND TEMPERATURE,   THE  VERTICAL BARS
                         INDICATE THE RANGE  OF THE DATA AS  A FUNCTION  OF HUMIDITY,

-------
                               APPENDIX A




                      MICROPHONE CALIBRATION METHOD






Introduction




     The microphone calibration procedure is performed in two stages.




The first stage employs a reciprocity calibration of the two condenser




microphones.  The second stage uses one of these microphones as a source




from which the response of the electret and ceramic microphones are ob-




tained.  The reciprocity method is used to calibrate the reference




(source) microphone at a particular temperature, humidity, and frequency.



The reciprocity technique requires three microphones.  One of these micro-




phones must be reversible, that is, it must perform according to certain




relationships between the acoustic pressure and velocity on the surface




of the microphone diaphragm and the current and voltage produced by the




microphone.  A third microphone is used as a source.  After the reference




microphone has been calibrated it is used as a source to generate a known




sound pressure from which the other microphone responses are determined.




The details and theory of how one performs these calibrations are presented




in Ref. 6.  The purpose of this appendix is to describe how HBS performed




these measurements.





Reciprocity Method




     Two different types of measurements are made in the NBS reciprocity




method of calibration:  voltage ratio and capacitance.  The voltage ratio




measurement is performed as shown in Fig. A-l.  With the switch in position




1 the oscillator excites the source microphone.  The receiving microphone,




acoustically coupled to the source through a 3-cm3 plane-wave coupler,
                                   A-l

-------
converts the acoustic signal into a voltage, which is filtered and  ampli-
fied.  The magnitude of the signal isiread on the meter and is denoted A.
The switch is then placed in position 2 disconnecting the oscillator from
the source microphone and connecting it to an attenuator calibrated in
hundredths of a decibel.  The output of the attenuator is connected across
the resistor R , which is in series with the receiving microphone.  The
              o
attenuator is varied until the meter indicates a value equal to A.  The
resulting attenuation reading, denoted A  (dB), gives the logarithm of
                                        3
the ratio between the open circuit voltage of the oscillator driving the
source microphone to the open circuit voltage of the receiving microphone.
Consequently the voltage ratio V., is given by V  = 10   a
                                K              R.
     Measurements of the capacitance of a microphone are made according
to the method developed by Koidan8 and is shown in Fig. A-2.  The source
and receiving microphone are placed in the coupler in the same manner as
in the voltage ratio measurement in order that the source microphone be
subjected to the same acoustic impedance.  When the two switches are in
position 1 a voltage is placed across the source microphone, which is in
series with a resistance R.  The reading of the meter, denoted Mj, that
is connected across the resistance R is recorded.  The two switches are
now placed in position 2 and the attenuator, which is calibrated in
hundredths of a decibel, is adjusted so that the meter again reads MI.
If the microphone is assumed to be purely capacitive with an impedance
at the given frequency that is much larger than R, the capacitance of
the microphone, C, is given by
                                   2irfRV3
                                   A-2
                                                                           (A-l)

-------
where f is the frequency of the oscillator  (Hz), V   e  10    c      and

A  is the attenuator reading in dB.

     The reciprocity calibration method used  three microphones,  denoted

//O, #1, and //2, and the two types of measurements discussed  above  to de-

termine the pressure response of two of them  (//I and  //2).  Microphone //O,

which will not be calibrated, is used as a  source.  With microphone  //O

as the source and microphone //I as the receiver the voltage  ratio  V   is

obtained.  With microphone #0 as the source and microphone #2 as the

receiver a voltage ration V  is obtained.   Lastly, with microphone //I

used as a source and microphone #2 as the receiver a  voltage ratio Vg is

obtained.  Using Eq. (A-l) this yields the  capacitance  of the microphone.

These three sets of measurements and some correction  factors described

below, will yield the pressure response of  microphones  //I and //2.

     From Section 4 of Ref. 6 the expressions  for the pressure  response

of microphones //I and #2, denoted r  and r2,  respectively, are
where
                        r  = K/G(f)^—    V/Pa                           (A-2)
                                    *v
                        r2 = K^GCO\—     V/Pa                           (A-3)
                                    n 2
                            G(f) = ^ A2(f ,fQ)                            (A-4)
                                    s
                                    sindrf/f )
                          *(f.f0) -  (rf/   '                              (A-5)
                                   A-3

-------
  In Eqs.  (A-2)  to (A-4)  f and R are defined as in Eq.  (A-l), P  is
                                                               s


 the ambient barometric pressure,  f   is  the  first  longitudinal  natural



 frequency of the coupler, and K is a constant.  For the setup  used  in



 this experiment the value of f  was  the frequency at which  the first



 maximum value of the output of the receiving microphone was obtained



 when the frequency was varied with a constant voltage applied  to  the



 source microphone.   The quantity K is a function of the volume of the



 coupler and the equivalent volumes of the microphones, adjustments  for



 the heat conduction effects at the walls of the cavity and the presence



 of the capillary tubes in the coupler,  and the ratio of the specific



 heats of air.   The  latter is virtually independent of temperature over



 the range considered, and while the volumes and capillary correction



 change very slightly with temperature,  they have been assumed to remain



 constant.   The  constant  K has been eliminated from the final results by



 referencing the r  to their respective  values at a given frequency and



 temperature.





 Comparison Method



     Comparison  calibrations  used  microphone #1  as the  source  and  the



 electret and ceramic microphones as the  receiving  ones.  The pressure



 response of these microphones is then determined from the  expression
                             K


                             — TV -     V/Pa
                                 s a
where fQ is the first longitudinal natural frequency of the coupler with



the microphone a as the receiver, V  is the voltage ratio,, K  is assumed
                                   a                        a

constant,
                                   A-4

-------
                                6 = ^                                    (A-7)









and r  is given by Eq. (A-2) and C by Eq. (A-l).  The constant K  is




eliminated by referring the pressure response to its response at a given




frequency and temperature.




     In arriving at Eqs. (A-l) through (A-7) several assumptions were made




regarding the effects of temperature on some of the constants in Eqs. (A-2),




(A-3) and (A-6).  Furthermore, there may be inaccuracies in the measure-




ments themselves.  The estimated maximum value of the errors introduced by




these two factors is tabulated in Table A-l which indicates that the total




maximum absolute error that exists in the results presented is probably




less than 0.2 dB.
                                   A-5

-------
                                TABLE A-l

               Estimated Magnitudes and  Sources of  Errors

                   in the Microphone Calibration Data
                 Source of Error

Assumptions that the following are independent of
temperature:

     1.  Volume of coupler

     2.  Coupler capillary corrections

     3.  Heat conduction correction


     4.  Equivalent volumes of microphones


Accuracy of measurement

     6.  Resonance frequency

     7.  Attenuator readings (combined)


     8.  Barometric pressure
      Estimated
    Maximum Error
        (dB)
        0.02

        0.02

0.06 dB @ 100 Hz
(less at higher freq.)

       Unknown
        0.02

  0.04 (reciprocity)
  0.05 (comparison)

        0.01
                                   A-6

-------
OSCILLATOR
                             COUPLER
   6 1

O	0

   0 2
                   PRECISION
                  ATTENUATOR
                                                       SOURCE MICROPHONE

                                                           CAPILLARY TUBE
                                                        RECEIVING
                                                       MICROPHONE
                                                 MICROPHONE
                                                PREAMPLIFIER
             FILTER
METER
FIGURE A-l.  SCHEMATIC OF EQUIPMENT REQUIRED FOR THE VOLTAGE RATIO
           MEASUREMENTS,

-------
00




ncpll 1 ATCI Q
UoU 1 LLA 1 U n







<
O. .,_

C






i 1
.A
* R
> 2
^M

PRECISION
ATTENUATOR





<
^
V
j
^M




•=

^
>
> \
— 4
i


rl^s
1*r
^
•• ^
\,
(ELE(

> 1
> 2


                                                     SOURCE MICROPHONE


                                                        -COUPLER


                                                     RECEIVING MICROPHONE
                                                 (ELECTRICALLY DISCONNECTED)
                                                                METER
         FIGURE A-2,  SCHEMATIC OF  EQUIPMENT REQUIRED FOR THE MEASUREMENT OF THE
                    CAPACITANCE OF THE SOURCE MICROPHONE,

-------
 NBS-1UA (REV. 7-73)
    U.S. OEPT. OF COMM.
   BIBLIOGRAPHIC DATA
         SHEET
1. PUBLICATION OR REPORT NO.

     NBSIR  76-1090
2. Gov't Accession
  No.
3. Recipient's Accession No.
 4. TITLE AND SUBTITLE
                              EPA 550/9-76-009
    Environmental Effects on Microphones of Various
    Constructions
                                               5. Publication Date

                                                 July 1975
                                                                       6. Performing Organization Code

                                                                           200.03
 7. AUTHOR(S)
    Gale  R.  Hruska.  Edward B. Magrab. William B. Penzes
 9. PERFORMING ORGANIZATION NAME AND ADDRESS
                                               8. Performing Organ. Report No.
             NATIONAL BUREAU OF STANDARDS
             DEPARTMENT OF COMMERCE
             WASHINGTON, D.C. 20234
                                                                       10. Project/Task/Work Unit No.
                                                                           2003110
                                              11. Contract/Grant No.
 2. Sponsoring Organization Name and Complete Address (Street, City, State, ZIP)
  Co-sponsored  by NBS and Office of Noise Abatement and Control
                           U.S.  Environmental Protection Agency
                           Washington,  D.C. 20460
                                              13. Type of Report & Period
                                                 Covered
                                              final,  March-June, 1975
                                              14. Sponsoring Agency Code
15. SUPPLEMENTARY NOTES
 6. ABSTRACT (A 200-word ot less factual summary of most significant information. If document includes a significant
   bibliography or literature survey, mention it here.)
 The  pressure^sensitivities  of two "1/2-inch" electret, two  "1-inch" ceramic, and  two
 back-vented  1-inch" condenser microphones were measured for  numerous  combinations  of
 temperature, percentage relative humidity, and frequency.   The  two condenser
 microphones were  calibrated  by the reciprocity technique at each combination of
 temperature, relative humidity and frequency.  The  condenser  microphones were then
 used as  calibrated sources  to determine  the pressure  sensitivities of  the  other
 microphones.  Insert voltage techniques  were used to  eliminate  the environmental
 effects  on the electronics.   It  was found  that the  back-vented  condenser microphones
 are  insensitive to changes in relative humidity.  At  frequencies considerably below
 their resonance frequencies  they exhibited only a very small  change in sensitivity
 with  temperature.   At frequencies closer to the resonance frequency the temperature
 coefficient increases approximately fourfold.  The  temperature  and humidity  coefficient
 for  the  electret  and ceramic  microphones could not  be  determined due to the  instability
 in their sensitivities which  produced changes that  were larger  than those  induced by
 the  temperatures  and humidities.
7. KEY WORDS (six to twelve entries; alphabetical order; capitalize only the first letter of the first key word unless a proper
  name; separated by semicolons)
 Calibration;  ceramic;  condenser;  electret; humidity; microphones; reciprocity
 sensitivity;  temperature.                                                        y'
8. AVAILABILITY           [£ Unlimited


  I  ' For Official Distribution. Do Not Release to NTIS


  I  ' Order From Sup. of Doc., U.S. Government Printing Office
     Washington, D.C. 20402, SD Cat. No. CH	
    | Order From National Technical Information Service (NTIS)
     Springfield, Virginia 22151
                                   19. SECURITY CLASS
                                      (THIS REPORT)
                                      UNCLASSIFIED
                                   20. SECURITY CLASS
                                      (THIS PAGE)

                                      UNCLASSIFIED
                          21. NO. OF PAGES
                                                             35-
                          22. Price
                           $4.00
                                                                                USCOWM-DC 29042-P74

-------