EPA-650/2-74-046-B




July  1974
                    Environmental  Protection  Technology Series
DE
                                              i:iii^^^^^

-------
                                        EPA-650/2-74-046-b
DEVELOPMENT OF  A GAS  LASER  SYSTEM

         TO  MEASURE  TRACE  GASES

                BY  LONG  PATH

         ABSORPTION  TECHNIQUES:
       VOLUME II - FIELD EVALUATION OF GAS LASER
             SYSTEM FOR  OZONE MONITORING
                     FINAL REPORT
                          by
               W. A. McClenny, F. W. Baity. Jr.,
             R. E. Baumgardner, Jr. , and R. A. Gray
                Chemistry and Physics Laboratory
              National Environmental Research Center
            Research Triangle Park , North Carolina 277]!
                         and
               R. J . Gillmeisfer and L. R. .Snowman
                  General Electric Company
                  Electronic System Division
                    100 Plastics Avenue
                 Pillsfield. Massachusetts 01201
                   Contract No. 68-02-0757
                     ROAP No. 26ACX
                  Program Element No. 1AA010

              EPA Project Officer: W. A.  McClenny
                      Prepared for
             OFFICE OF RESEARCH AND DEVELOPMENT
            U.S. ENVIRONMENTAL PROTECTION AGENCY
                  WASHINGTON, D.C. 20460
                        July 1974

-------
This ruport has been reviewed by the Environmental Protection Agency
and approved for publication.  Approval does not signify that the
contents necessarily reflect the views and policies of the Agency,
nor does mention of trade names or commercial products constitute
endorsement or recommendation for use.
                                 11

-------
                      TABLE OF CONTENTS






                                                           Paze No.
A. INTRODUCTION	 1





B. DESCRIPTION OF THE SYSTEM  	,	3




C. SYSTEM CALIBRATION 	 10




D. PATH-POINT MONITOR COMPARISON METHODOLOGY	14




E. FIELD MEASUREMENT RESULTS  	31




F. CONCLUSIONS	,	 43




G. REFERENCES                                              4G

-------
                          LIST OF ILLUSTRATIONS
                                                                      Page No.
Figure 1 - ILAMS Block Diagram                                          4
Figure 2 - "V" Laser Optical Layout                                       G
Figure 3 - Data Collection and Reduction System                            9
Figure 4 - Response to Ozone in Calibration Cell                           12
Figure 5 - Space-Time Plane for Measurement Path                        17
Figure 6 - Time-Delayed Correlation Between Ozone Point Monitors         21
Figure 7 - Ambient Ozone Variations About Mean Ozone Concentrations       23
Figure 8 - Open-Path Measurement Site                                   24
Figure 9 - Super-Imposed Recorder-Traces                               27
Figure 10 - Super-Imposed Recorder-Traces (Space)                       29
Figure 11 - Super-Imposed Recorder-Traces (Time)                        30
Figure 12 - ILAMS and Auxiliary Equipment in Trailer                      32
Figure 13 - A View of the Optical Path Over Which Ozone was Measured      33
Figure 14 - Comparison by Use of Procedure 3                             36
Figure 15 - System Performance Effect from Beam Movement on the         38
               Retroreflector
Figure 16 - Comparative Path/Moving Point Monitor Ozone Data             41
Figure 17 - Point and Path Monitor Comparison                            44

                          LIST OF TABLES
Table I Syracuse N. Y. Ozone Concentrations                                2
Table II Characterization of Ozone Variability Along Measurement Path       26
Table Til Point Monitor-Path Monitor Comparison Data by Procedure 1       34
Table IV Shift of Beam Position on Relroreflector with Time of Day as        42
        Indicated by Focusing Lens Adjustment
                                     IV

-------
 A.  INTRODUCTION
        The Final Report of EPA Contract 68-02-0757, Development of a Gas Laser
 System to measure Trace Gases by Long Path Absorption Techniques, consists of
 two (2) volumes:
        I.  Gas Laser System Modifications for Ozone Monitoring
       II.   Field Evaluation of Gas Laser System for Ozone Monitoring
        The work reported here  stems from development activity begun in 1966 at
 General Electric's Electronics Laboratory.  Under this contract, a breadboard
 laser long path monitor called ILAMS (Infrared Laser Atmospheric Monitoring
 System) was modified to improve its sensitivity as indicated by previous field
 experience.  System parameters were selected to optimize system performance
 for ozone monitoring.  A field evaluation of the modified system was conducted.
        Following completion of the system design investigations and hardware
 modifications described  in Volume I of the  Final Report, the laser path monitor
 called ILAMS was set up and operated over a range (one way distance) of . 67 kilo-
 meter.  Field tests were performed in cooperation with EPA personnel.  The
 object was to evaluate the monitoring capability of the breadboard ILAMS by
 comparing its ambient air measurement of ozone concentrations with data from
 chemiluminescence point monitors operated near the system's optical path.  Table
 I shows representative chemiluminescence  ozone monitor data from Syracuse,  New
 York, where the measurements were made.
       This portion (Volume II) of the Final Report,  is a joint EPA-General Electric
Company effort.  In the sections which follow, the system is described.   Calibration
of the system with a multipass cell and point/path monitor comparison methodology
are discussed,  field test  results are presented and conclusions are given.

-------
Month + Year


Jan 1973
Feb 1973
Mar 1970
Apr 1973
May 1973
June 1973
July 1973
/ ug 1 973
Sept 1973
Oct 1973
Nov 1973
Dec 1 973
Jan 1974
Monitoring Station
No. 1
Average of
Daily Averages
10
15
14
20
21
27
35
32
17
11
9
10
9
Range of Maximum
Hourly Averages
5-35
0-40
1 -48
14 -59
17-75
16 -87
20 - 125
30 - 100
9-93
15 - 57
9-49
2-32
4 - 32
Downtown
Average of
Daily Averages
8
11
14
19
18
*
39
34
no dot
no dot
Range of Maximum
Hourly Averages
5-29
5 -57
6-96
12 - 108
10 - 76
24 - 124
30 - 132
22 - 132
a
a
no data
no dot
a
no data
Data from New York State Continuous Air Monitoring System

Number of daily averages (18) insufficient,  per New York State data processing
criteria,  for taking monthly average.  Average for 18 days is 36 ppb.
Table I  -  Syracuse, N. Y. Ozone Concentrations During Recent Months
           (in Parts per Billion )

-------
 B.  DESCRIPTION OF THE SYSTEM


       The gas laser system used in this evaluation is a breadboard instrument.

 It was constructed by General Electric Company under an internally funded program

 and its performance evaluated in rural and urban atmospheres under EPA Contract

 EHSD 71-8.  Under this  contract the system was modified as described in Volume I of

 this document.  The system operates in the middle region of the infrared spectrum

 and identifies atmospheric constituents by absorption spectroscopy.   It measures

 average pollutant concentrations (total burden) over its optical path.  Laser operation

 is at relatively low, safe power densities of  .001 to .01 watts/cm2 in the spectral

 region where the eye does not transmit.


       Figure 1 is  a block diagram of ILAMS.  The output power from the  laser is

 directed to a 50 percent  beamsplitter via a 1 mm spatial filter (cleanup aperture).

 The energy reflected from the beamsplitter is focused down to a 0.1 mm  aperture

 that serves as an  attenuator.  Behind this aperture is the reference energy detector.

 The transmitted power through the beamsplitter goes to a germanium lens which

 focuses the energy near  the focal point of an  off-axis parabolic mirror, and the

 expanded,  nearly  collimated beam is transmitted to the retroreflector.  The return

 energy from the retroreflector retraces the path through the beam-expanding parabolic

 mirror and the germanium lens to the beamsplitter.  The return energy reflected

 from the beamsplitter Is collected by a germanium lens doublet and focused on the

 signal detector.  Preamplifiers are mounted directly behind the signal and reference

 detectors.  The preamp outputs  go to the signal processor.  The detectors used in

 the system are thermistor bolometers,  operating at ambient temperature (uncooled),

having a  characteristic flat response across the middle infrared spectral  region.


      The attenuation of laser energy at many wavelengths produces absorption

patterns  which are used to separate pollutant absorption from spectral interferences
                         1 *?   1 C
in the signal processor.  C  O     laser lines in the P branch of the 00°1 - 02°0

-------
                         POLLUTED ATMOSPHERE
                                                "7
                                          RETROREFLECTOR
                              V /
                                LENS
 REFERENCE
 DETECTOR
                ATTENUATOR
LINEAR
WEIGHTS
DOUBLET
   LENS
                              SIGNAL
                              DETECTOR
                                  SPATIAL
                                   FILTER
                                   SPECTRALLY
                                   SCANNING-
                                   LASER
     DIGITAL
SIGNAL PROCESSOR
                       Figure 1 .  ILAMS BLOCK DIAGRAM

-------
 transition lie in the same spectral region as ozone absorption so that coincidences
 between the two exist.  Several of the absorption coefficients recently measured
 by EPA personnel are listed in Volume I.  Prior to the actual monitoring of ambient
 ozone, a set of four wavelengths, optimized for ozone detection in the presence of
 other atmospheric attenuators, was chosen by an  optimization procedure.   This
 procedure and the calculation of linear weights is described in Volume I.  During the
 course of the evaluation some  changes in wavelengths and weights occurred.  The
 final wavelengths were 9.2938, 9.5039,  9.5862 and 10.5321 microns.  For these
 wavelengths the linear weights were respectively  -0.4215,  1.0000, -0.5820 and 0.0034.
        The spectrally scanning laser itself includes a high gain  CO  flowing gas
 laser as the radiation source and a wavelength selection mechanism,  which
 periodically (50 Hz) scans through a series of four laser wavelengths.  A typical output
 scan consists of laser pulses of 2. 5 msec, duration separated by time intervals of the
 same duration during which laser action is interrupted and detector null signals are
 recorded.  The laser optical configuration is shown in Figure 2.
        The laser cavity consists of a "V" shaped plasma tube and an external  spectral
 tuner,  A relatively long laser  cavity is used for sufficient gain to overcome the
 losses inherent in the spectral  tuner and to obtain lasing action on a large number of
 spectral lines. A beam travels through the plasma tube with aid  of a  mirror at the
 point of the "V".  Leaving the tube through a germanium Brewster window,  the beam is
 directed by mirrors through an  iris (for  mode control) and onto a 105 lines/mm diffraction
 grating, which disperses the beam spectrally and spatially.   The four wavelengths of
 interest are then  relayed through holes in the chopper wheel to the four end mirrors of
the laser cavity.  These holes in the chopper are so located that, as the wheel turns,
only one wavelength at a time is permitted to pass through to the  end mirrors.  The four
end or wavelength selection  mirrors are adjusted so that the beams are directed back on
themselves through the laser cavity.  In this way,  selected laser wavelengths are
transmitted sequentially.

-------
Sr-..tia!
 Hirer
                               Coupling Mi noi
                                    Beam to/from
                                     Col limator
                                                                                                                    Concave
                                                                                                                    Mirror
                                                Signal  Bolometer
                                              Reference         Concave Mirror
                                              Bolometer
IRIS Stop          /             S

            4  End Mirrors   Chopper Wheel
Brewster
Window
                                        Figure 2.  "V" LASER OPTICAL LAYOUT

-------
       The mini-computer signal processor includes a general purpose (stored

program)  mini-computer and appropriate interface electronics.  The collection

and reduction of data is entirely under computer, i.e. , program control; results

are displayed on simple displays incorporated in the equipment, and on an optional

teletype, which need not be used (or even be connected) during field or test range

exercise of the system.

           The use of the stored program control and data reduction means:

       •   changes in system design, or variations in data reduction algorithms,
           may be accommodated without alteration of the data  collection  or re-
           duction hardware; only changes in the control  program will be  required.

       •   modification of signal processor parameters such as number of wave-
           lengths (up to 8), gate locations,  system response time, weighting
           factors, etc. , do not even require software changes, these parameters
           are expediently entered by the teletype  input.

       •   the precision of data processing may be made as accurate as desired;
           similarly the impact of imprecise calculations may be assessed by direct
           simulation for purposes of evaluating future low cost special purpose
           instruments.

       •   additional data,  e.g., environmental conditions,  time,  date, signal
           variability,  laser parameters, etc. , may be measured and recorded
           without modification of or addition to the existing system  hardware.

       •   the performance of one  or more data processing and display  systems
           can be directly analyzed,  e.g., data from several ozone monitors
           could be crosscorrelated  and  recorded.

-------
       The data collection and reduction system is sketched in Figure 3.  A Digital
Equipment Corporation PDF 11/05 is used for the central processor.  The data
collection and reduction equipment in Figure 3 consists of three major subsystems:
Interface Subsystem
       This subsystem includes an 8 input analog  signal multiplexer, which is
followed by a sample-ancl-hold amplifier and an  analog-to-digital converter at
10-bit precision.  (The analysis path detector preamplifier output is connected to
one multiplexer input, the reference path to a second multiplexer input, the re-
maining  6 are available for sensing other voltage levels of interest).  Additional
L-.I=> system elements include an AGC attenuator,  a wheel  position counter and
demultiplexer/storage capability for analog data displays like the meters shown
in Figure 3.
Central Processor Subsystem
       The central processor and its own control panel form this subsystem.   Power
supplies  for this equipment are contained within the CPU cabinet proper.  The
central processor control panel ordinarily is disabled during operation.
Program Input and Data Logging Subsystem
       A Teletype Corporation ASR-33 teletype  with appropriate interface circuits
constitutes this subsystem. As indicated,  it plays two roles. First, it permits
entry (ordinarily via paper tape) of the control program.  Second, it permits detailed
reporting of directly measured quantities,  or derived (computed) quantities.
       The central processor is designed so that programs  stored in its core memory
may be caused  to remain intact during periods of no primary power.  This option
is exercised, so that once a control program has been entered in the CPU, it
need not  be  reentered until there is a need to change it, regardless of whether the
CPU remains energized or not.  The control program is designed so that it will
run properly regardless of whether the teletype  is  connected or not.  Thus the
teletype unit is an  optional data display device, not an essential component of the
system once the control program has been entered.

-------
Analysis Path
Amplifier Output

Reference Path
Amplifier Output

Chopper  Wheel
Encoder Timing H
Signals
                       Interface
Multiplexer,
Sample and
Hold, A/D
AGC
Attenuators,
D/A Converter
Storage Registers
                      (S)
                      Meters
                                                 Central Processor
                                                 PDP 1 1/05
                                 CPU
Teletype and Paper Tape Reader
(Teletype Not Required for Equipment
Operation; May Be Removed After
Control  Program Has Been Loaded
     Program Input and
     Data  Logging
                    Data Display
                                              Figure 3.  DATA COLLECTION AND REDUCTION SYSTEM

-------
 C.  SYSTEM CALIBRATION
        In day-to-day operation, system zero was set by comparing concentration data
 on the teletype printout with the readings of a portable chemiluminescent ozone point
 monitor operated along the system's optical path. An appropriate offset was entered
 in the signal  processor Parameter Table so that the value on the printout corresponded
 to the average concentration indicated on the point monitor strip chart record.
        For ozone, and pollutants with similar diurnal variations, a simple alternative
 procedure can be followed.  Typically, ozone concentrations were zero from sun-
 set until  shortly after sunrise.  During this period, the system  can be zei-oed by
 offsetting to zero the log of transmission at each wavelength.  Adding this fixed
 offset in  the signal processor is equivalent to introducing an optical  filter that com-
 pensates  for the fixed absorption pattern introduced by the system optics.
        Prior to field operation, the system was calibrated using a multipass (White)
 cell* into which known ozone concentrations were introduced.  In the experimental
 setup, the uncoilimated laser beam was mirror-directed into the entrance port of
 the multipass cell.  Cell  mirrors were so aligned that the beam was redirected
 back on itself, rather than leaving the  cell  through the exit port. Two small ozone
 generators employing UV illumination  were arranged in series to produce a 1.6 ppm
 ozone concentration in  air flowing  into the cell at four liters per minute.  This
 corresponds to S.O percent absorption at the strongly absorbing wavelength, A   9.505^,
 and to an  equivalent ozone burden along the measurement path of 50 ppb (parts per
 billion).   Ozone concentrations  at the entrance and exit of the calibration cell were
 monitored with a portable chemiluminescence ozone monitor** (Analytical Instrument
 Development,  Inc.) which had been previously calibrated by comparison with an
 ozone generator, which was  in turn calibrated by the neutral buffered potassium iodide
 technique.  No appreciable ozone loss was observed.
* Fabricated by EPA
       contribution to field tests
                                         10

-------
       For another run under the same conditions, the ILAMS was used as the sensor
of ozone concentration buildup in the cell.  In the system, the teletype printout is the
data display device.  On the printout each group of digits (five at the time of the
experiment but now four) represents the sum of the weighted natural logarithms (in)
of the return signal to  reference signal voltage ratios for each of the transmitted
wavelengths.  This sum is the system response.  It is proportional to a. CL,  the
exponent of the Beer's Law equation,  where a is the absorption coefficient, C the
concentration and L the system's optical path length.  Since ap CL  is dimensionless,
the units of a are the reciprocal product of those used for C and L as will be  seen
below.
       For this experiment,  the system's four wavelengths  (9.305, 9.504, 9.586 and
10. 532 microns) were  weighted for ozone as the target gas and the sum appeared as
the left hand group of digits on the printout.   The weights were,  respectively,
-.4131, 1.0000, -.5869 and zero.   As the ozone concentration in the multipass cell
increased to an equilibrium value, system response was  recorded  on the teletype
printout.  At equilibrium the  reading on the teletype printout was 00064, corresponding
to an a CL of . 128.  This compares with an a CL of . 119 derived from multipass cell
parameters and the point monitor measurement of ozone concentration as  shown in the
following calculations.   Figure 4 shows how ILAMS response and ozone concentration
measurements relate,  in terms of a CL, as cell ozone concentration increases with
time to an equilibrium  value.  Intermediate points on the plot are determined in the
same manner  as shown below for equilibrium values.
                 a CL Calculation from Multipass Cell Parameters
Measured 0  absorption coefficients,  a, at transmitted wavelengths are as follows:
           O
                                 01  (atm   cm
9.305
9.504
9.586
10.532
0
12.659
.69183
0
                                       11

-------
   .130

   .120

   .110

_,  .100-1
u
•6
 ,  .090

!£  .080-1
O

|  .0701

2  .060-
z
O  .050-
O
Q_
Z  .040-

   .030-


   .020-


   .010-
                                        /
                        IDEAL CURVE
       0   ,010 .020 .030 .040  .050 .060 .070 .080 .090 .100 .110 .120 J30

                          ILAMS RESPONSE -  c* CL
        Figure 4.  RESPONSE TO OZONE IN CALIBRATION CELL
                                      (ILAMS VS. POINT MONITOR)
                                     12

-------
The resultant a  ,  the sum of weighted absorption coefficients at the four


wavelengths is:



       a-  = 4131 x 0 + 1. 0000 x '12. 659 - . 5869 x . 69183 + 0 x 0
        r


          = 12.253 atm"  cm"



The multipass cell was aligned to produce 42 passes through a 5 foot (1.524 meter)


cell length.  For an assumed 1.52 ppm equilibrium On concentration in the cell


(obtained from previous point monitor measurements employing the same procedure)


we have:

                                 ft

       a CL = 12.253 x 1.52 x 10~  x 42 x  1.524 x 100 = .1192
        r
                                       13

-------
  D.   PATH-POINT MONITOR COMPARISON METHODOLOGY




          There are several methods by which comparisons between point




  monitors and path monitors can be accomplished.   For each of these methods




•  it  is  important to assess the comparison'accuracy and to insure that




  tolerances  inherent to the method are not attributed to disagreement




  between  the  monitors.   The differences between methods correspond  to the




  ways in  which  point monitors  are  used to  obtain  values of the  ambient  trace




  gas concentration  for  use  in  the  comparison and  to  the statistical  treat-




  ment of  this data.   The sampling  procedures for  use  with  point  monitors



  are the  following:





          1.  Recording  the  real time signal from  a point monitor as  it




             is moved along the measurement path.




         2.  Simultaneously taking several bag samples along the path




             and measuring each with a point monitor.




         3.   Monitoring  the signals from stationary point monitors placed




             along the measurement  path.




         4.   Filling a sample container while  moving along the measurement




             path and measuring the path integrated sample with  a suitable



             point  monitor.





         In what  follows, these  various sampling procedures and  the  corres-




ponding point monitor,  path monitor comparison methods  are discussed




with special emphasis on their  applicability for  ozone.




        In Procedure  1  a point  monitor  is  moved with  a  constant  velocity, V,




along the measurement path, thereby describing a  trajectory on the space-




time graph of Figure 5,  i.e., a straight line given by S = VT.  This




procedure like others using

-------
 point monitors does not give complete space-time coverage.  It can be thought of as
 providing a number, N, of concentration measurements, A., along the measurement
 path.  A specific A. corresponds to a spatial and temporal increment and a
 coordinate (S., T.).  The average value, n  , obtained for the trace gas density during
 a transit time, T ,  is   £  A./N.    Often the point monitor data can be obtained  as  an
                       i
 analog signal recorded on a chart recorder. In this case
                 "!-';    -
                          .X = 0                                         (1)

where A is now a continuously recorded value for the trace gas concentration and
corresponds to position along a chart trace.  If conditions remain the same along
the path until several traverses are made, the values of n  associated with a set of
traverses (point monitor measurement sequence) form the statistical basis for a
determination of the  precision with which the true mean value of path averaged con-
centrations can be stated. Specifically, if a sample mean n  is obtained for the
individual values of n ,  then the 95% confidence interval n  ± A   includes the true
mean value for 95% of all possible measurement sequences.  The value of  A
is given by the expression (References 1 and 2).
                            l        .05  ll                         (2

where s  denotes the standard deviation for a set of traverses, k  denotes the
number of traverses in a set, and t nr. is a tabulated statistical function which
                                . 05
approaches 1. 96 for large k .
                                        15

-------
       Once the confidence interval has been defined, a formal comparison can be
made between the value n  and the average value,  n*, of path monitor readings during
the point monitor measurement seouence.  Experimentally, the number of path
monitor readings is usually much greater than k  while the standard deviation for
path monitor readings is approximately the same.   Thus,  the value of A  associated
with the confidence interval for the path monitor is much less than in the point
monitor case and the  value of n* closely approximates the true mean.  For a given
comparison seouence, if n* lies within the interval n  + A  95% of the time, then
a statement that the two monitors agree can be made.  Procedure  1 was used in a
comparison sequence discussed in a later section of this report.
       Procedure 2 involves establishing  a number of stationary positions along the
path at which bag samples are taken.  For a bag filling time, T , the space-time
coverage consists of vertical channels in Figure 5  corresponding to complete time
coverage during T  at the fixed sampling positions.  The spatial interval A£ over
which the point monitor readings are representative,  i.e. ,  the  width of the vertical
channels in Figure 5, is determined by the main features in the spatial gradients
which exist along the path. In the case of ozone,  these main features are caused  by
topological proximity conditions, e.g. , trees or buildings which provide surfaces on
which ozone is destroyed, thereby constituting an ozone "sink"  and by NO source
proximity such as automobile traffic near or across the measurement path.  Ozone
disappearance occurs when NO concentrations comparable to ambient ozone con-
centrations are present.  In this case the  rapid reaction between these two gases
(Reference 3) can cause ozone changes to  occur within a few seconds.  Both of these
causes of spatial  gradients are dependent on the residence time of ozone/NO near
sink/ source regions and therefore change as a result of wind speed and direction.
                                        16

-------
            SPACE
3.   ^FAOK-Tr-I". PL-VS";-. FOR
      l-"'AS'''HP!F.::T  PAT!'

-------
 For Procedure 2, placement of the sampling sites must be such that spatial
 gradients are revealed.
        Ozone concentrations must not be changed in the sampling and measurement
 procedure.   If NO levels are high the gas phase reaction of NO  and O  inside sample
                                                                  O
 bags can reduce the O  level drastically.  Ozone can also  be destroyed on the walls
 of the sampling bag.  Typical decay curves for 100 liter Tcdlar and Mylar bags
 indicate that ozone decay rates can be limited to less than S9c per hour if the  sample
 bags are conditioned with ozone prior to use and if the bags are filled to insure
 high volume to surface  ratios.
        If the sample site selection is appropriate and the integrity of bag samples is
 assured, an accurate point-path comparison can be made.   Each bag sample
 measurement represents a mean ozone concentration during T .  If T. is short com-
 pared with temporal variations in ozone concentrations,  several samples should be
 taken at each point.  The average of a  set of simultaneous  samples taken •along the path
 can be treated analogously to the path-averaged reading  obtained during  a traverse
 in Procedure 1.  Each average defines a mean value n  so that a confidence interval
 n9 _L   A ^  can  be defined based on several values of n.  n  replaces n   and   A
 replaces   A] in the statement on point-path comparison.  If T  is long compared with
 temporal ozone  variations, the readings obtained from the sample bags closely
 approximate the true mean values at the sample sites and the sum of readings from
one set of sample bags can be compared directly to an n* established during T .
       Using Procedure 3, a close approximation to the  true average trace gas con-
centration at some point along the measurement path can be obtained by monitoring
for a time period long compared to that typical of trace gas variations.   This
experimental technique  is illustrated for O  later in this  section.  Once these values
are obtained, they arc simply averaged and compared with ;m n* established during
                                          18

-------
 the same time period.  Just as in Procedure 2, stationary site locations must be
 adequate to accurately represent any spatial gradients.  This is obviously not the
 case when only one or two point monitors are used along a long path near which
 sources and sinks of the trace gas exist. Thus, the usefulness of Procedure 3 for
 point-path comparisons depends on the choice of the path and prior  knowledge of the
 temporal and spatial variability of the target gas.
        Procedure 4 is useful in those cases when a portable point monitor for the
 target pollutant does not exist and when the path cannot be adequately characterized
 by sampling at a number of fixed sites.  A sample container is  filled as the path is
 traversed so that different portions of the  path  are equally weighted.  The container
 is carried to a suitable point monitor and a single reading representing the target gas
 average along the path is obtained.  Target pollutant loss while in the  sample container
 must again be considered.   However,  in the case of relatively inert tracer gases or a
 primary pollutant such  as CO, this loss can be negligible.   The value of the reading
 obtained for each traverse, n , is treated like n  and the comparison technique is
 the same as that using Procedure 1.
        Measurements were made to determine the short term temporal/spatial
 variability of ozone.  Other than diurnal variations in ozone concentration and long
term  trends established by NO or O  sources, e.g., the effect of traffic or general
 cloud cover patterns, variations within periods  of several  minutes occur which reflect
localized variations  in target gas concentration  in the air mass  being carried across
measurement locations.  These localized variations arise  due both to disruption of the
NO, NO2, O  , and ultraviolet radiation steady state (Reference 4) by changes in the
ultraviolet throughput to ground level due to segregated clouds and to pockets  of gases,
e.g.,  NO or olefinic hydrocarbons, with which ozone  reacts.  These effects give rise
to short term structure in the ozone concentration in both space and time.
                                          19

-------
Two ozone point monitors (Bendix Model 8002;  were used in field tests to establish
the variability in ambient ozone concentrations.  These monitors detect ozone by its
gas phase chemiluminescence with ethylene and have minimum detectable limits of
less than 5.0  ppb.  One monitor was placed on a stationary platform and a second one
was mounted on an electric  car and moved to different points along the  measurement
path.  Figure 6 shows chart  recorder traces of the two monitors as a function of time.
Both monitors were stationary and they were separated by 42 meters.   Variation in
ozone concentrations are the result of wind conditions,  of NO source and O.  sink
                                                                       O
proximity, and of cloud cover patterns, i.e.,  reduction of the short wavelength
                    O
radiation ( ^  < 4000A) necessary to photolyze NO .   Air mass transport from one
monitor  site to the other is  obvious from the time delay correlation of signal features
on the two chart recordings.  Correlations of this type vary depending on the
projection of the wind vector along the path joining the two point monitors.  Intake
ports for the monitors were established at a height of six  feet above the ground.
These ports were open Teflon tubes of 8 mm inside diameter and approximately 100
cm long.  Flow rate into the instrument was approximately 1.0 liters per minute
and the time constant for the instrument was set at 1.0  seconds.  With these  experimental
conditions the ozone concentrations monitored by the  instruments were  essentially
real time.
       In order to characterize temporal variations  for one point monitor given
meterological and source/sink proximity conditions,  a method of data handling was
developed.  A single chart recording is characterized by two quantities, an average
concentration  c and a measure of the temporal variability  of ozone.  Data from the
chart  recording was digitized at six second intervals  and c was determined by averaging
the set of digitized data values.  The digitized readings were fitted by a linear re-
gression analysis over a short time period, typically ten minutes.  The differences
                                     20

-------
    50
   40
o 30
i-
cc
I—
LU
O
o 20
 cr>
O
   10
          "i    i	1	1—i	1	1	1	1	1	1	1	1—i—i	1	1—i	r
                    CORRELATION (DELAYED)
CORRELATION
  (DELAYED)
          I     I     I     I     I     I     I     I     I     I     I     I     I     I     I
                                                I	I
                                                       10
                                                    TIME, min
                                      15
20

-------
 between the linear approximation and actual recorded values were obtained and the
 number N of these differences within given intervals,    AC, of the straightline
 value were tabulated.  The distribution of N with  Ac was determined to be approx-
 imately a Gaussian distribution  and hence was determined uniquely by a mean value
 and standard deviation.  A typical distribution with a superimposed Gaussian is shown
 in Figure 7.  As expected the standard  deviation changes in response to variations
 in source/sink proximity conditions and in meteorological parameters such as wind
 speed and direction.  Values of  the standard deviation could be used to obtain a value
 of the true mean ozone concentration  during the period of measurement.  Typically,
 100 values were determined so that a 95% confidence interval of "£"±0.196s was
 defined.  In most cases c is determined within one or two ppb and can be used for
 point-path comparison as suggested in the discussion of Procedure 3.
        Of the procedures discussed for point/path comparisons, the first and a
 variation of the third were used  in field tests described in the next section.  The
 major advantage in using these two procedures is that minimal ozone  loss occurs
 due to ozone confinement prior to measurement.  A secondary advantage is the
 requirement of only one person to perform that part of the comparison involving the
 point monitor.  Field measurements along a measurement path of 0.67 Ion have
 indicated some of the characteristic variations in ozone concentrations to be
 expected.  Both a hand-carried Model 560 portable ozone monitor made by Analytical
 Instrument Development  Inc.,  and a Bendix Model 8002 ozone monitor were used to
 obtain real time ozone concentrations  by moving  along the measurement path.  The
 Bendix instrument was placed  on an electric car  and powered by batteries with DC
 inverters to render it portable.  A stationary monitoring  station was  established near
one end of the measurement path as shown in Figure 8. One ozone monitor and a
Climet wind speed and direction  indicator were placed at  the stationary site.  A path
along which the point monitors could be moved was established from the laser source
                                        22

-------
                                   Viy:.;?   ,; .   ''1S!.Yi 0?.0.\r VAT AT TO/'S
o
Q_
U_
O

O
                                                                                      MEAN = 0.001
                                                                       STANDARD DEVIATION = 4.05
                                            DIFFERENCE AC (ppb)

-------
pE\:-p..\TH  '••T>\STJRP 'FNT  S


(G.F-.. Sy?ACIiSF,,  N'.Y.)
                                                O     O
                                                  FUEL OIL

-------
to its retro-reflector.  Runs with the moving point monitors prior to




actual comparison trials indicated that the concentration variability




of ozone was strongly dependent on the wind direction and the constancy




of cloud cover conditions.  Arteries of traffic existed to the west and




south of the stationary monitoring station while a forested residential




area was to the north and east.  During periods of two to three hours




after 2:00 p.m. (DST) on days with consistent cloud cover, almost constant




ozone concentrations existed.  Consistent light wind conditions from the




northeast were optimum for low ozone concentration variability.




        In preparation for comparison tests, a number of simulated point




monitor comparison sequences were run.  The objectives of these tests




were to establish the spatial gradients and the variability of ozone




concentrations along the selected measurement path.  The distribution of




the average concentration values (n-^ values) during the individual




traverses was assumed to be normal and the comparison was treated as




specified in Procedure 1.  Table II summarizes the point monitor data.




Test 1 of Table II corresponds to the information obtained from six path




traverses.   A reproduction of the recorder chart outputs from these tra-




verses are shown superimposed in Figure 9.  Each trace was obtained by




hand-carrying the AID ozone monitor along the measurement path of 0.67




km in a time of approximately eight minutes.  A mean ozone concentration




of 68.2 ppb is shown in Figure 9 along with dotted lines indicating the




95% confidence interval.   For each traverse the recorder chart data




was digitized to give approximately 100 values for the ozone concentration.







                               25

-------
Test
1





2*





3




nl
69.9
73.6
64.2
66.9
68.5
66.2
62.5
63.5
64.3
65.5
64.0
63.2
68.9
70.4
70.8
69.9
70.7
nl
68.2





63.8





70.1




i
Al
3.4





1.1





1.0




stl
4.9
5.4
6.4
7.1
7.0
10.0
1.0
2.2
2.4
1.8
2.4
2.2
2.2
2.7
2.1
2.4
2.2
SH
6.8





2.0





2.3




Mode
Moving





Moving





Stationary




Figure
9





10





11




     * Three of the traverses were made with a  1.0 second time constant
       and the remaining three with a 10.0 second time constant.
Table II   Characterization of Ozone Variability along Measurement Path
                                26

-------
   100
    80
    60

-------
The standard deviation, S   , of these values varied appreciably from




traverse to traverse as seen in Table II.  This variation is caused by




the contribution to the concentration average of the large decreases




evident in Figure 9.  Decreases of this type were caused by the inter-




action of ozone with pockets of air having high concentrations of NO.




This is an example of the practical limitation to accuracy of a compar-




ison test in which the ambient atmospheric conditions are not sufficiently




equivalent.  However, in other simulated comparisons, during another day,




the values of the standard deviations were approximately equal, indi-




cating constant atmospheric conditions.   Figure 10 and the corresponding




treatment of the data, i.e., Test 2 of Table II illustrate this point.




For Test 2, a Bendix ozone monitor with response characteristics equi-




valent to the AID instrument was used.  The confidence interval was




approximately three times smaller than that of Test 1.   Thus, ambient




conditions could be chosen during which the accuracy of point-path




comparison sequences would be high.  Specifically, the atmospheric condi-




tions prevalent during Test 1,  related mainly to wind direction, result




in a wider confidence interval, + A , about a sample mean than do those




prevalent during Test 2.




        Figure 11 shows superimposed recorder chart traces of five




consecutive time intervals taken under the same atmospheric conditions




as the traces of Figure 10.   The data was recorded on a second Bendix




ozone monitor at a stationary position along the path.   Near equality




of the standard deviations associated with the stationary (Test 3) and




and moving (Test 2)point monitors demonstrate the equivalence of temporal




variability associated with a stationary monitor and the temporal-spatial




variability associated with a moving point monitor.



                                     28

-------
   100
                                        BENDIX MOVING MONITOR


                                             SIX TRAVERSES
    80
    60
o

o
o
co
O
40
    20
     0
       LASER
                   140
280               420

   DISTANCE, meters
560
RETRO
                    ripfi:"re
                                 r'PPFFT) PF.r^P^1"" -  TRACK? (SF.^CF!^ 11  ."rjOPFP "1°7T"

-------
    100
    80
    60

-------
 E.  FIELD MEASUREMENT RESULTS
        Field tests employing ILAMS over a 0. 67 km path  were conducted during the
 period October through December of 1973.  The laser beam was positioned between
 1  and 2 meters above ground level and pointed along a horizontal path inside the plant
 boundaries of General Electric Company in Syracuse, New York.  The
 ILAMS was housed in a small,  air-conditioned trailer at one end of the path.  In-
 dependent  suspension of the laser and associated optics was provided in order to
 maintain the pointing accuracy  of the laser since small changes of the order of a few
 thousandths of an inch at the source correspond to displacements on the order of
 inches at a remotely positioned retroreflector.  The beam was transmitted to a two-
 mirror type retroreflector, consisting of an aluminized parabolic reflector of 122 cm
 focal length and a small secondary mirror (1.0 cm diameter) at the focal point, was
                                                      2
 positioned  at the other end of the path. An area of 600 cm  (30 cm x 20 cm) was inter-
 cepted by the retroreflector.
       Figure 12 is a picture of the system and auxiliary equipment in the trailer.
 Figure 13 is a view from the laser of the optical path over which ozone was measured.
        Actual path/point monitor comparison tests were made in two ways. In the first,
 Procedure  1, described in the previous section,  was used.  The individual averages,
n  were obtained by measuring  the area under chart recorder traces.  Two sets of six
such values are  shown in Tests  1 and 2 of Table III.  Ozone variability was similar to
that shown  in Figure 9. Some variation in individual values of n  occurred due to changing
conditions along the entire path. As specified in Procedure 1,5, and  A  were
determined. The average path  monitor reading, n*,  during each set of six traverses
was obtained by averaging a digital printout from the signal processing unit of the
monitor (4  second update)  which consisted of approximately 600 determinations of the
path averaged ozone concentration.  Under ideal test conditions n* should fall within
                                       31

-------
Lrl
tsj
                  Figure 12. \LAMS AND AUXILLIARY EQUIPMENT IN TRAILER - SYRACUSE, N.Y

-------
Figure 13. A VIEW OF THE OPTICAL PATH OVER WHICH OZONE WAS MEASURED.
                                 33

-------
Test
1





2





nl
31.6
32.6
33.8
34.1
37.7
35.1
49.8
56.5
43.7
43.2
44.5
47.0
nl
34.2





47.5





Al
2.2





5.3





n*
31.3





49.3





Table III    Point Monitor-path Monitor Comparison Data by Procedure 1
                                  34

-------
 the confidence interval of n 95% of the time.   For the data presented in Table III,
 Test 1, n* = 31.3 ppb falls just outside the confidence interval 34.2 + 2.2 ppb.   For
 Table 2, Test 2, n* = 49.3 ppb falls well within the confidence interval 47.5 + 5.3 ppb.
        To obtain a better measure of the ability of one monitor to track the other,
 comparison sequences during which ozone concentration gradients occur were deemed
 appropriate.  In this case individual values of n were compared directly with n*,
 i. e. , the average value of the ozone concentration along the path  as determined  by one
 traverse was compared with the average value'of the path monitor readings during
 the time of traverse.  If concentration gradients do not exist,  such as is the case
 shown in Figure 9, each point along the measurement path is equivalent.  By dividing
 the chart recording of a single traverse into a number of segments and assigning a
 concentration value to each,  a data base for a statistical treatment to determine the
 confidence interval for n is established.  Equation 2 applies where now  A  defines the
 confidence interval about  n, s denotes the standard deviation in a  sample of y values
 for the  concentration.  For y^ 100, t    * 1. 96 and using values  of s from Table II
                                   • UD                                        *
 Test 3,   A is less than 1.0 ppb.  Even with s as high as 10.0 ppb, the confidence
 interval is only 4 ppb.
       In actual practice, the recorder chart trace for each traverse was measured
 with a planimeter and the  value obtained was compared directly with the average of the
 path monitor readings  during 1^  Figure 14 shows a comparison of this type extending
 from noon until sunset and includes the points presented in Table III as  Test 2.  In
 this comparison sequence good agreement is obtained although the disparity between
 path monitor and point monitor readings for several pairs of points is greater than
would be expected for a common measurement path, under equivalent atmospheric
 conditions.   Review of  the chart recording  show  that departures  from
 equivalence  occur for comparisons which show  the largest disparity  .  In
 this case path monitor  and point monitor may  see quite different concentration
 profiles.  This  is  illustrated in Figure 1 by the  placement of areas labeled
 A. and B.  Significant  decreases in ozone concentration  in  these areas of the
 space-time plane  will cause measurement-disparities.
                                        35

-------

-------
        In the period following the October comparative path/point monitor tests, the
 ILAMS system was operated extensively.  In the course of this work a number of areas
 were investigated.  Studies of system alignment, beam position on retroieflector and
 distance between retro primary and secondary,  were made and optimal conditions
 established.  These studies utilized alternative  entries in the Parameter Table that
 were preserved on paper tape for reference and future use.  Laser performance was
 improved  by adjustments and modifications.  The teletype printout format was
 changed from 4 groups of five digits to eight groups of four digits.  Other program
 changes to improve system performance were implemented.  Minor linear weighting
 changes were made to correct previous errors.  Signal processor sensitivity to
 elevated ambient temperature in the trailer was found to be caused by a defective
 component.  System operation ended with a series of path/point monitor comparative
 tests 26 November through 4 December.
       Again,  as in the October tests, system performance effects from movement of
 the laser beam on the retroreflector were noted. Figure 15 shows the rapid change
 on the teletype printout as the beam moves off the retro. Data is  from 3 December tests.
 Printout speed is twelve lines per minute.  Each count on the left  hand, ozone column,
 represents 2.44 ppb of O^  That is, 0010 indicates 24. 4 parts per billion of ozone.
 As can be seen from the figure,  in about  two minutes, counts went from 0013 to 0001,
 a change of 29 ppb.
       It is possible to compensate for beam motion through movement of the focus-
 ing lens in the system.  This lens focuses the laser beam on the entrance aperture
 of the collimator, an off-axis parabola optical system which serves as the breadboard
 ILAMS transmit-receive optic. It is mounted in  a fixture on the laser channel.  The
 fixture is oriented so that the lens  can be moved  along the laser beam axially,
vertically (V) or horizontally  (H) in relation to  it. Consequently, V and H axis
adjustment of the focusing lens can be used to move and position the system's transmit
beam on the retro.
                                        37

-------
 -''.(•'•1M  -PI? P. 5
 '""" 13  - f P V (:
 (.• . r- {•/"  - ?• f/; ]  !
 P'f'.!./:  -"PPS    ffT'/i  -?:«.!.''    '"'r'C<:.


                       :f-l"/i  -•?••'•• .17    PP^ii

                       •ipr/t.  -i/:\! 7    !-'.p;it

• P P.!  ?  - F. ? P. • 5    f-' P!••• /;  - P p. j. 7    r/' r- £ 7
"CO.!?.  -P!P-P5    P'P-R'j  -P-P17    P'P'J-6

•!''f:U3  -{VPP.5
•«?ej.r?  -PPPS
• (.j; CM 2  - P ('• p 4
                       p!'P-3  -P'Cj.6    P'P.1.5
 •P'P'J?  -PPP/J    "(?•!"•"}  -pTr    *  - '. J    ". •_ ".• »i*    '. , .'.».-    '    / *»'

 •ff.1.2  -PPP'3
 •P'0'.i..l  -PPiP.3
 •?p.i.!  -; (';:{;.    f?-f.,--:-  -f.I.OC-
- P-'-1! .'•;    '•'(• '• "f  ••'-  i /•''
- '•' •'  (•"!',    '.• '"• /: V  - '' I '•'' i
- P-'••:•• .'5    (.' (''/: •:  - f/- 1 /.; t
- r/,. :.- ••    (/(.•/.• f".  - f": i .':'••
-1.'(«'.••'.?    P'///;  -!-:.i^rv



• PfM •''    '-'I' -!••'-'•  -("'.I t' !
-PP..I. i     ?•("'"*;£  -C:.J :''  - T-  1.7 V
•ppr:  -r-pj/s   -r'.i'ei. f/o-c:     CppP-
-r'"f:'P:rj    pfp-^i  -f'^r'-s  -pf-r'o     pf-2-.1
-K(.'?i2    i'-C'-P1.^  -^.^^5  -f-p^     t>'(-H:'.
- ft ft •/. r;    r/ R f. c;  _ r> i/ />. c,  _«-,. r- r-.;,     .-, r. •- c,
- (V. f-' r^ ;„;    f- f.: p. K  _ t; p f, j-,  _ ,;: /> ,-;,:;     f, r .;..'(.

-P !''("'[    C'!/i.V£  - f» f-\ (.'•' f,  -r-.^-'f- .'••'     ?•'•• C'.'lA

- f  i.'. !'•' !    r-1 r.p r... i^  _ p- ^ ft ('..  _ (i p: f,. -i     f n '* '•


  fff''~'i'~    (i (/ <:•;e;  _(;;..•."(';!,:   _/v iv f.-'/     (/''~/-•:-.
  ^•^•;v;;-    i'-i:. f:.-1.  _f^rti:   .. (v r..-/; y     §.;'"(;. *'     • • •'••' ";i>
                                                  - P C 3 i   - P. P p17  - <•.- i v-:<
                                                  - PCI'!•    P ff'P 6  - f; i ^ 7
                                                    P P I P    f". Cf 06  - fJ-! '. ^
                                                    P'p^7     pp?^;  -r-VV:>
                                                    P P. f' v     P P 3 5  - *"'' I ;.• 7
                                                  • P.' C- ('•' /';     p P- ? r-:  - f• J;"; 7

                                                  - P f'; f- •"'.     p « p £  - f ] «; •;•
                                                  - ff' f-  !:;     ''f'ff..'.'.  - f.- ', •"•.t
                                                  -P !''•-=!     Pp.l>^  -r.i.1,-..^
                                                  •r*p:."-3   -wf-T-j   -(.-;:;;;,
                                                   pr;-f^^   .. '/(.•!/ i   ^ ••; i r. •;
                                                   f.r.*r  -f,,.,.;.   -;-i7!
                                                   '."••(.'} •!    ('<:.*<:••:   ->. \ |- /;

                                                   r- r;i j '..    (.,./ (;.:•;   _ p j r:;. v



                                                   !.'T-3i    r~.•••]'•-;   -•('•'1,^9
Figure  15.  SYSTEM PERFORMANCE EFFECT  FROM BEAM MOVEMENT
              ON THE RETROREFLECTOR
                                            38

-------
        Like the teletype printout, the return (analytic) signal amplitude reflects
 beam position on the retro.  Maximum amplitude indicates the beam is centrally
 located on the retroreflector.  With return signal amplitude oscilloscope-displayed,
 centering the beam on the retro can be done by V and H axis adjustment of the
 focussing lens. This approach was used to check beam position and center the beam
 on the retro.   Signal amplitude variation with V and H movement was an indication
 of beam location relative to the retro center.  Thus, maximizing or "peaking" the
 signal on the oscilloscope with V and H adjust would center the beam.  When centering
 the beam, it was decided that final V and H settings would be the mid-point between
 the micrometer screw settings where the return signal rose and fell significantly  as
 the beam was moved  across the retro by the motion of axes adjustment screws.
        A series of comparative path-point monitor field tests in the period 26 November
 through 4 December duplicated some of the results obtained in October and provided
 additional information on the system's performance.  Comparative data was collected
 in the same manner as that employed in the earlier tests.  Retroreflector  alignment
 checks, secondary  mirror position relative to primary,  took place 28 November.  The
 secondary was moved .025 Inches further away from the primary.  Also a weight  change was
 made on the ozone channel to reflect an earlier  shift in  A , from the R14  line to the
 R16 line of the (00  1 - 02  0) transition. In microns, the wavelength shift was from
 9.3054 to 9.2938.   It was made to avoid water vapor Interference noted after the original
 wavelength selection process was completed.
       The weather during this period in November and early December was quite variable
 with snow and rain occurring. The system was  operative in all but extreme weather
 conditions, even when the  retroreflector site was not visible from the trailer.  December
 3rd was an unusually warm winter day.  The high temperature was 58° F, a near-
 record for Syracuse.  It was bright and sunny, reminiscent of important data-taking
 days in late October.  The laser was turned on at 0650.  The system was balanced at
 0830.   Around 0900  a  steady negative drift in concentration-related  counts on the
teletype printout prompted freo^ient checks of beam position on the retroreflector
throughout the day.
                                        39

-------
       The comparative data for this clay is shown in Figure 1G.  Noteworthy is the
genera] absence of large excursions in path monitor relative to point monitor data
noted in October data during mornings and late afternoons, fit  was felt that adjust-
ment of beam position in the period 1345 to 1430 could have been more frequent to
reduce the excursion at 142.S on the figure). Table  IV  shows beam position shift with
time of day in terms of focussing lens adjustments made to maximize  the return
signal.  Xote that  H axis adjustment is constant (within operator reading error
tolerance for an awkward location) while V axis setting rises and falls with time of
clay and coincidentallv,  the effect of localized  atmospheric heating.  This represents
about a seven centimeter beam excursion on the retro.  Similar data for 23 October
shows the laser beam raised 15 centimeters on the retro reflector.  As the retro
diameter was 20 centimeters, this motion could be expected to have considerable effect
on the system.  The time  of day variation in beam position gives rise to speculation
that beam motion at the retro is largely attributable to atmospheric looming.
                                     40

-------
0
                                                                               *  PATH MONITOR
                                                                                  MOVING POINT MONITOI

                                                                                ESTIMATED FROM
                                                                                LATER READING
            900
1000
                                  1100
                      1200
                  TIME OF DAY
                                                                                        1600
            Figure 16.  COMPARATIVE PATH/MOVING POINT MONITOR OZONE DATA - 3 DEC 1973

-------
                          3 December 1973
Time
Start*
0815
0928
1006
1022
1135
1259
1410
1440
1513
1540
1615
V
379.6
376.5
378.0
378.9
H
308.5
308.4
308.5
308.4
no change
379.6
308.4
no change
379.3
379.0
378.5
378.0
377.4
308.5
308.5
308.5
308.5
308.5
                Laser was turned on at 0650.  Initial V and H settings
                were those made at 1520  30 November.
Table IV     Shift of Beam Position on Retroreflector with Time of Day
             as  Indicated by Focusing Lens Adjustment
                                 42

-------
 F.  CONCLUSIONS
        An example of point monitor and path monitor comparisons is shown in
 Figure 17.  It includes the afternoon data of Figure 14 with the morning data
 taken on 23 October.   The erratic behavior of the T.LAMS during the morning
 hours  has not been definitely identified,  but is believed to be related to the
 alignment of the laser beam at the retroreflector.  Spurious absorptions result
 whenever the relationship between return and reference signals arc changed by
 other than known atmospheric attenuators.  In this case the spurious absorption
 coincided with drift of the transmitted laser beam away from the retroreflector
 caused by atmospheric looming as the ground warmed up and cooled down  during
 the day to produce vertical temperature  gradients in the system's optical path.
        The accuracy  with which measurement of ozone  concentrations could be
 made was limited by optical noise related to atmospheric turbulence and thermal
 gradients.   The transmit and receive optics produced spectra.!  attenuation  because
 the diameter of the far field diffraction pattern of the laser beam is proportional
 to wavelength. The retroreflector is an  aperture stop that acts upon these spatial
 variations to produce spectral attenuation.  Hence, when the transmitted beam was
 focused on the retroreflector, one would expect  a spectral return in the presence
 of atmospheric looming, which is precisely what we observed.  It was shown possible
 to produce a similar error by deliberately aiming the beam slightly off the edge of
 the retroreflector.  The experiments with defocusing  the beam to alleviate this
problem, however,  insulted in an increase in the observed optical noise and an
 increase in measurement error.  The aggravation of the noise condition with de-
focusing indicated that spectral non-uniformities were still present in the near field
 in spite of the spatial  filter (clean up aperture) in the  laser transmitter. Direct
                                       43

-------
      o POINT MONITOR
      o PATH MONITOR
                                           SYSTEM
                                          REZEROED
0800
1000
          1200          1400          1600

         TIME, hours (23 Oct.)

Figure J.7.   POriT  ANT)  PAT1! MONITOR COMPARISON
1800

-------
 measurements of the intensity pattern of the near field showed that there  is indeed
 some "clutter" in the near field beam and this has been tentatively attributed to the
 optical surfaces of the laser transmitter both preceding and following the clean up
 aperture.
        The signal output of this laser detection system is proportional to the log
 of the ratio of the signal  returned from the retroreflector to the signal on the
 reference detector.  Since it is the  relative signal behavior of the two detectors
 to which  the output is sensitive, it is equally possible that drift in the optics pre-
 ceding the reference detector is also partially responsible  for the measured error
 in the output.  However,  because the major drift error occurred  in the mornings
 on clear  clays following clear nights, the  strong change in humidity and ground
 temperature during the morning hours are  considered significant.
       Water vapor is fairly well understood spectral interference, but ground fog
 and liquid water adhering to participates  (haze)  is not.  In addition, unknown spectral
 interferences may have entered the  beam to offset the system's zero baseline.  (A
 method called factor analysis,  to analyze and compensate for unknown spectral
 effects is discussed in Volume I).
       The effect of looming on the  laser beam was noted on many occasions.  Direct
 measurement of the effect has been  reported in the previous section.  Reaiming the
 laser by moving the focussing lens preceding the beam expander to correct for the
 looming only appeared to  compensate for  about 50% of the measurement error associated
 with this time of day phenomenon.  The considerable drift in the laser system output
 signal which has been tentatively attributed  to looming was greatest during the mornings
 on clear days when the wind velocity was low.  In these periods, the RMS error
 between the ILAMS and the point monitor ozone measurements was approximately 28
 parts per billion.  At other times, the system was relatively stable and the RMS error
 was about 6 parts per billion. It is expected that the effects of atmospheric looming on
 the signal are still the dominant source of error even during stable operation and that
correcting non-uniformities in the transmitted laser beam will reduce the  system
 error and improve its accuracy.
                                           45

-------
                                REFERENCES
1.  G.W. Snedecor, Statistical Methods, The Iowa Suite College Press, Ames,  Iowa.

2.  B. E.  Saltzman, Journal of the Air Pollution Control Association, _22_(No. 2),
    99, February 1972.

3.  J.A. Ghormley, R. L. Ellsworth, and C. J. Hochanaclel, J. Phys. Chem. ,  77,
    1341, 1973.

4.  D.H.  Stedman, E.E. Daby, F.  Stuhl, and H. Niki, Journal of the Air Pollution
    Control Association, _22_(No.  4), 260, April 1972.

5.  Final Report:  Field Study on Application of Laser Coincidence Absorption
    Measurement Techniques, Contract EHSD 71-8,  February 1972,  prepared
    for the Environmental Protection Agency by General Electric Electronics
    Laboratory, Syracuse, N. Y.  The system used in the comparison tests is a
    modified form of the system described in this final report.  Modifications
    include those specified in Reference 5.
                                   46

-------
4. TITLE AND SUBTITLE
  Development of a Gas Laser System  to  Measure Trace
  Gases  by  Long Path Absorption Techniques   Volume II
  Field  Evaluation of Gas Laser System  for  Ozone Mom'tori
J. f'ErU'GRMING ORGANIZATION NAME AND ADDRESS
  Joint  Report:   Environmental Protection  Agency, Researc[i
  Triangle  Park,  North Carolina  27711  and
  General Electric Company, Electronic  Systems Division,
  Pittsfield,  Massachusetts  01201
                                  TECHNICAL REPORT DATA
                           (I'leasc rcail [nuruciiciHS on t/ic rcvcrsi1 before c
 . REPORT NO.
 EPA-650/2-74-046-b
                5 REPORT DATE
                  July 1974
                6. PERFORMING ORGANIZATION CODE
                la.
                                                           8. PERFORMING ORGANIZATION REPORT NO.
 W. A.  McClenny,  F.  W Baity, Jr.,  R.  E.  Baumgardner, Jr.
 R. A.  Gray,  EPA, R.  J. Gillmeister  and  L.  R.  Snowman, G
12. SPONSORING AGENCY NAME AND ADDRESS
 Environmental Protection Agency
 National Environmental  Research Center
 Chemistry and Physics  Laboratory
 Research Triangle  Park,  North Carolina
27711
                                                            RECIPIENT'S ACCESSIOC*NO.
                 _  __     __
                "lO. PROGRAM F LEMEM NO.
                     1AA010
                11. CONTRACT'GRANr NO.

                     68-02-0757
                 13. TYPb Oi REPORT AND PERIOD COvi nbD
                 	Final_ _   _  	
                 14. SPONSORING AGENCY CODE
15 SUPPLEMENTARY NOTES
 Joint Field Study by G. E.  and  EPA of Open-Path Monitor
1G. ABSTRACT
 Ambient ozone measurements  in  real  time using an open-path  monitor are described.
 These studies establish  the sensitivity of an open-path  monitor, based on  transmissiv
 ity  measurements of C02  laser  lines, at <_ 5 ppb and  validate the values obtained during
 real-time monitoring of  ambient  ozone by establishing  and  using a methodology  for
 the  comparison of point  monitor  readings and open-path monitor readings over a  common
 path.
17.
                                KEY WORDS AND DOCUMENT ANALYSIS
                  DESCRIPTORS
  Lasers
  Atomospheric Absorption
  Ozone
  Air Pollution Monitoring
                                              b.IDENTIFIERS/OPEN ENDED TERMS
     I LAMS
     Methodology for  Point
     Monitor, Path Monitor
     Comparisons
c. COSATI I'ickl/Oruup


  1705
'>'-:. _M5TRIBUTIO'Ni STATEMENT
  Release Unlimited
    19. SECURITY CLASS (This Report)
      Unclassified
21. NO. OF PAGES
   51
                                              20. SECURITY CLASS (This pugt!)
                                                Unclassified
                                                                         22. PRICE
EPA Form 2220-1 (9-73)
                                             47

-------
                                                           INSTRUCTIONS

     1.    REPORT NUMBER
          Insert the EPA report number as it appears on the cover of the publication.

     2.    LEAVE BLANK

     3.    RECIPIENTS ACCESSION NUMBER
          Reserved for use by each report  recipient.

     4.   TITLE AND SUBTITLE
          Title should indicate clearly and  briefly the subject coverage of the report, and be ili\pla\ ul prominent!'.. Set subtitle, if used, in smaller
         type or otherwise subordinate it  to main title. When a report is prepared in more than one volume, repeat the primary title, .uld \ulume
         number and include subtitle for  the specific title.

     5.   REPORT DATE
         l-.ach repoit shall carry a date indicating at  least month and year.  Indicate  the basis on which it was selected fc.v., ilatc of'/.u/i,'. /
         approval, date of preparation, etc./.

    6.   PERFORMING ORGANIZATION CODE
         Leave blank.

    7.   AUTHOR(S)
         Give namelsl in conventional order (John R. Doc, .1. Robert Doc. etc.). List author's affiliation if it differs from tlu' pcrformmi' oiuani-
         '/ation.

    8.   PERFORMING ORGANIZATION REPORT NUMBER
         Insert if performing organization wishes to  assign this number.
     i
    9.   PERFORMING ORGANIZATION NAME AND ADDRESS
         Give name, street, city, state, and ZIP code. List no more than two levels of an organi/alional hircarchy.

    10.   PROGRAM  ELEMENT NUMBER
         L'se the program element number under which the  report was prepared. Subordinate numbers may be included in paionthescs.

    11.   CONTRACT/GRANT NUMBER
         Insert contract or grant number under which report was prepared.

    12.   SPONSORING AGENCY NAME AND ADDRESS
         Include ZIP code.

    13.   TYPE OF REPORT AND PERIOD COVERED
         Indicate interim final, etc., and if applicable, dates covered.

    14.   SPONSORING AGENCY CODE
         Leave blank.

    15.   SUPPLEMENTARY NpTES
         Enter information not included elsewhere but useful, such as:  Prepared in cooperation with, Translation of  Presented at conference of
         lo be published in, Supersedes, Supplements, etc.

    16.   ABSTRACT
         Include a brief (200 words or less) factual summary of the most significant  information contained in the report If the report contains  i
        significant bibliography or literature survey, mention it here.

    17.   KEY WORDS AND DOCUMENT ANALYSIS
                                                is of Engineering and Scientific Terms the proper
                                                                                            ataloging

        (b) IDENTIFIERS AND OPEN-ENDED TERMS - Use identifiers for project names, code names, equipment designators etc   Use open-
        ended terms written in descriptor form for those subjects for which no descriptor exists.

        (i) COS ATI HELD GROUP - Field and group assignments are to be taken  from the 1965 COSATI Subject Category List  Since the ma-
        jority ol documents are mullidisciplmary  in  nature, (he Primary Field/Group assignment!.-*) will be specific discipline area of human
        endeavor, or  type of physical object.  The application(s) will be cross-referenced with secondary Field/Group assiunmcim thai will follow
        the primary postmg(s).                                                                '                -       "

   18.  DISTRIBUTION STATEMENT
        Denote reusability to the public  or limitation for reasons other than security for example "Release Unlimited " Cite anv ",viihhililv to
        the public, with address and price.                                                                   '        •  '    '

   19. & 20.  SECURITY CLASSIFICATION
        DO NOT submit classified reports to the National Technical Information service.

   21.  NUMBER OF PAGES
        Insert  the tola! number of pages, including this one and unnumbered pages,  but exclude distribution list, if any.

   22.  PRICE
    ;    Insert the price set by the National Technical Information Service or (be Government Printing Office, if known.
(a) DESCRIPTORS - Select from the Thesaurus of Engineering and Scientific Terms the proper authorized terms that identify the nnior
concept ol the research and are sufficiently specific and precise to be used as index entries for cataloging.
:PA Forrr 2220-1 (9-73) (Reverse)

                                                              48

-------