EPA-650/2-73-030



October 1973
ENVIRONMENTAL PROTECTION TECHNOLOGY  SERIES







                                                                           ill;
                                                    £-^ Jh'XwZ* I»A* jL&'jL**jL&jl4£'lt jjjl'Z£'l'frbQL'^i-L-^-£^'l'^'i'~S.-'. •'.•'.•'.•'•'

-------
                                      EPA-650/2-73-030
              INSTRUMENT
TO  MONITOR CH4/ CO, AND  C02
          IN AUTO EXHAUST
                      by

            D. E. BurchandJ. D. Pembrook

                 Philco-Ford Corp.
                Aeronutronic Division
                    Ford Road
            Newport Beach, California 92663


               Contract No. 68-02-0587
             Program Element No. 1AA010


        EPA Project Officers: J. Sigsby and F. Black

            Chemistry and Physics Laboratory
          National Environmental Research Center
           Research Triangle Park, N. C. 27711
                   Prepared for

         OFFICE OF RESEARCH AND DEVELOPMENT
        U.S. ENVIRONMENTAL PROTECTION AGENCY
              WASHINGTON, B.C. 20460

                   October 1973

-------
This report has been reviewed by the Environmental Protection Agency and




approved for publication.  Approval does not signify that the contents



necessarily reflect the views and policies of the Agency, nor does




mention of trade names or commercial products constitute endorsement




or recommendation for use.
                                 11

-------
                               ABSTRACT
An infrared analyzer employing gas cell correlation techniques and
thermoelectrically cooled photodetectors has been designed and con-
structed to measure the concentrations of methane, carbon monoxide, and
carbon dioxide in automotive exhausts.  The dynamic range from the mini-
mum detectable concentration to the maximum concentration that can be
measured accurately for each gas is:  methane, 0.3 ppm to 10,000 ppm,
carbon monoxide, 0.5 ppm to 10,000 ppm, and carbon dioxide, 10 ppm to
250,000 ppm.  The wide dynamic range is made possible by employing two
different sample cell lengths for each gas.  The concentrations of the
three gases are measured simultaneously and independently.  Discrimi-
nation against other gases in the automotive exhaust is very good.

This report was submitted in fulfillment of contract number 68-02-0587
by Philco-Ford Corporation, Aeronutronic Division, under the sponsor-
ship of the Environmental Protection Agency.  Work was completed as of
August 1973.
                                  iii

-------
                              CONTENTS




                                                                   Page



Abstract




List  of Figures                                                     v




Acknowledgements                                                    vi




Sections




I      Conclusions and Recommendations                              1




II     Introduction                                                 3




III    Optics                                                       12




IV     Electronics                                                  21




V      Test Data  .                                                  29




VI     Glossary of Symbols                                          36



VII    Patent Information                                           39
                                   iv

-------
                                FIGURES

No.                                                                Page

 1       Block diagram of a single gas analyzer                     5

 2       Rotating gas correlation cells                             6

 3       Calculated spectral plots of transmittance for a
         model spectrum similar to a portion of a CO band           8

 4       Calculated spectral plots of transmittance
         demonstrating the response of the gas correlation
         system to a gas other than the one to be detected          10

 5       Schematic diagram of the optics assembly                   13

 6       Expanded view of slit assembly S2 or S3                    18

 7       Power supply unit above a single electronics unit          22

 8       Schematic illustration of a single channel of the
         photodetector electronics within the optics assembly       24

 9       CO detection sensitivity                                   30

 10      CH^ detection sensitivity                                  31

 11      C02 detection sensitivity                                  32

-------
                            ACKNOWLEDGEMENTS
We wish to give special recognition to the ingenuity of Mr. F. J. Gates,
a member of our research team.  He devoted much of his own time to the
development of the instrument and to the testing of its performance.

We appreciate the cooperation and the candor of EPA technical personnel,
particularly Mr. F. M. Black in discussions about the instrument and
its specifications.  The EPA personnel at Raleigh-Durham, North Carolina
also furnished us with an analyzed sample of lead-free gasoline for
testing the instrument.
                                  vi

-------
                               SECTION I

                     CONCLUSIONS AND RECOMMENDATIONS


This program has been completed successfully.  An instrument has been
designed, constructed, and tested.  All the major design criteria were
met.  The instrument can measure CH4, CO, and C02 simultaneously and
independently.

The measurement range includes gas concentrations that vary approximately
from those of room air to those of raw automotive exhaust.  The system
has very good discrimination against other gases that occur in auto-
motive exhausts.  Water vapor is the only gas that interferes enough to
consider a correction factor to the measurements of CO, ClU, and COo.
If the exhaust gas sample contained 3% water vapor, the maximum error
resulting from it would be 0.6 ppm (part-per-million) of CO,. 2.3 ppm of CH4,
and 30 ppm of COo.  The concentration of water vapor in the sample can
easily be determined, or estimated, to within +0.5%.  Thus, the maximum
error, in the measurements due to water vapor can be reduced with a cor-
rection factor to 0.1 ppm for CO, 0.4 ppm for CH^, and 5 ppm for C02.

The large dynamic range of the instrument necessitates the use of two
different sample cell lengths for each gas.  Cells of two lengths are
active at all times for the CO and CH^ systems.  The operator selects
between the two photodetector outputs from the two cells with a switch.
Electronic automatic ranging could be incorporated into the instrument.

In the 2 system, only one of the two sample cells is in the optical
path at a time.  A bi-stable mechanical cell position device allows the
operator to select between and reposition the cells in a few seconds.
Only one photodetector is required for this system.

The spectral passbands of the CO and CH/ channels of this instrument are
defined with a moderately complex grating-filter assembly.  This assembly
allowed the passbands to be adjusted and selected for good sensitivities
to CO and City and very good discrimination against other gases.  If the
requirements on discrimination were relaxed, the optics assembly could be
simplified by employing other methods of defining the spectral passbands.
                                  -1-

-------
 If  only  one  sample cell  length were required at a time for the CO and
 CH4 systems,  then the optical throughput of these systems could be
 increased.   This would result in an increase in the signal-to-noise
 levels of these channels.  Further, only one photodetector would be
 required for  each channel.  A single cell device would have a dynamic
 range that is an order of magnitude less than that of the present instru-
 ment.  The C02 system demonstrates one technique for mechanically switch-
 ing between two sample cells.  Similar techniques could allow switching
 between three or more cells of different lengths for extended dynamic
 ranges.

 The photodetector output signals of each channel are processed independ-
 ently by three separate electronics units that are almost identical to
 each other.   This modular construction technique allowed thorough testing
 of the first unit before design approval and construction of the subse-
quent units.   If one electronics unit becomes inoperable, either of the
remaining units can be substituted.  The total volume occupied by the
electronics  could easily be halved by repackaging the three electronic
units into a single cabinet.   If this were done,  the parts count could
also be reduced by about 25 percent by using a single power supply and
reference amplifier.
                                  -2-

-------
                               SECTION II

                              INTRODUCTION
PERFORMANCE

This section presents sufficient detail for a general understanding of
how the exhaust gas analyzer functions.  The operational characteristics
are summarized.  This is followed by a general description of the optics
and the electronics.  The spectral principles of operation are described
in the last portion of this section.  Subsequent sections present
detailed descriptions of the instrument, its operation, and its calibra-
tion and performance.

The exhaust gas analyzer provides for the simultaneous measurement of CO,
CH^, and C02-  The concentrations of the individual gases are analyzed
with a three-channel gas correlation spectrometer.  The output signal is
proportional to the absorber thickness of the sample gas, which is the
product of the concentration of the gas species being measured and the
sample cell length under the fixed conditions of room temperature and
one atmosphere pressure.  Each channel employs two sample cells of dif-
ferent lengths to extend the useful measurement range.  The CO and the
CH^ systems have common sample cells.  Either system may use either or
both cells at any time.  All of the photodetectors are thermoelectrically
cooled.

With a 1 second time constant on the most sensitive ranges, the minimum
detectable gas concentrations are approximately 0.3 ppm  (parts-per-million)
of CH4, 0.5 ppm of CO, and 10 ppm (0.001%) of C02.  The maximum gas
concentrations that can be measured reliably are 10,000 ppm of CH^,
10,000 ppm of CO, and 250,000 ppm (25%) of CO?.  The gas correlation sys-
tem offers very good discrimination between the gases being measured and
other gases that occur in automotive exhaust.  The system is also insen-
sitive to particulates in the sample, dirty windows, and changes in
source brightness.
                                   -3-

-------
 EQUIPMENT DESCRIPTION

 All three channels of the instrument  are  similar.   Many  of  the  component
 parts are shared.   For clarity,  the  following description is  limited  to
 a single channel.   Figure 1 is  a block diagram of  any  channel of  the
 device.   Radiant energy from an infrared  source is  imaged on  a  photo-
 detector through a chopper,  the  rotating  gas  correlation cells, one of
 two sample cells,  and a narrow  bandpass filter.  Optical components
 between  the source and the photodetector  were selected and  arranged to
 give good stability along with  a large throughput.  The  energy  is chopped
 at 360 Hz (Hertz)  by an opaque,  12-bladed,  chopper  deposited  on the entrance
 window of the  rotating correlation cells.   Figure 2 shows an  exploded
 view of  the correlation cells.   One of the  cells GC2,  is filled with  a
 non-absorbing  gas  and has a  neutral attenuator deposited on the window
 of this  cell.   The other cell, GCl, is filled with gas of the species
 being detected.   Sufficient  gas  is placed in  this cell to cause equal
 attenuation by both cells over  the spectral bandpass defined  by the fil-
 ter in front of the photodetector.  The narrow bandpass  filter  assembly
 may be a simple interference filter or a more complex  monochromator.  The
 complexity of  this assembly  is determined by  the requirements of sensi-
 tivity to the  gas  species of interest  and the rejection  of, or  insensi-
 tivity to,  other  gases  that  may  be present  in the unknown sample gas.
 The sample  gas  flows  continuously through the sample cell to minimize
 errors that  could  be  caused  by adsorption or  desorption  of  gas  on the
 sample cell walls.

 The gas-filled  correlation cells rotate at  1800 rpm, which  has  the effect
 of switching the optical path between  the correlation  cells at 30 Hz.
 When no  sample  is  present in the sample cell,  the photodetector output
 consists  of  a  360  Hz  carrier signal due to  the  360 Hz  chopper.  When  the
 sample cell  contains  gas  of  the  same species  as the gas  in  the correlation
 cell,  the photodetector  output consists of  a  360 Hz carrier modulated at
 30 Hz.   The  amount of  30  Hz  modulation  is proportional to the absorber
 thickness of the detected gas present within  the sample cell.  Thirty Hz
 modulation  to  the  carrier occurs only when  the  sample gas absorbs energy
 within the  system's  spectral bandpass.  Further, for modulation to occur,
 the  absorption must  correlate to some  extent  with the  absorption charac-
 teristics of the gas within  the gas correlation cell.  Placing a neutral
 attenuator within  the  sample cell would not cause 30 Hz modulation since
both halves  of  the  correlation cell cycle would be affected equally.

The carrier may exhibit 30 Hz modulation when no sample is present if
one correlation cell window  collects more dust than the other or if the
temperature of the correlation cell changes.  Since these events are
certain to occur, no attempt was made  to achieve perfect balance between
the cells.  Instead, an electronic 30 Hz modulator is employed to cancel
any residual modulation that occurs.
                                  -4-

-------


















































£


Amplifier

i


Phase Shifter


^

Adjustable
Phase Shifter


































































— »


















f
360 Hz Tuned
Amplifier

J
Gain Controlled
Amplifier

4

Modulation Balance

|






AGC Control Amplifier




















1














































^













Range Switch

4
Calibrate Pot

|

30 Hz Tuned
Signal Amplifier
J

30 Hz Synchronous

j

Adjustable Time

Constant Amplifier



Meter and
Recorder Output




ELECTRONICS UNIT





















'Figure 1.   Block diagram of a  single  gas  analyzer.

-------
-WINDOW
CHOPPER
ATTENUATOR
                                                                     -WINDOW
Figure 2.  Rotating gas correlation cells.  Windows are antireflection overcoated sapphire.

-------
SPECTRAL RESPONSE

The spectral sensitivity of the gas correlation spectrometer is set by
the narrow bandpass filter and by the absorption characteristics of the
gas in the gas correlation cell.  The CO detection system is used to
illustrate the operating principles.  The use of the carrier at a fre-
quency, fc, of 360 Hz does not affect the spectral sensitivity of the
instrument.  If the carrier chopper were removed from the system shown
in figure 1, the detector output would occur at the frequency, fa, of
30 Hz.  During one half the cycle GC2 would be in the beam; during the
other half cycle the gas filled correlation cell, GCl, would be in the
beam.

Figure 3 contains several calculated plots of transmittance that can be
used to explain the spectroscopic principles of detection.  We have
assumed certain gas sample parameters and have chosen a band model
consisting of two absorption lines with spacings, strengths, and half-
widths typical of the stronger lines of the fundamental CO band.  The
half-widths at 1 atmosphere pressure are 0.08 cm"1; strength = 2rr; and
the lines are separated by 4 cm"1.  Correlation cell GCl is assumed to
contain 0.2 atm cm of CO at a total pressure of 1 atm.  Its transmittance
spectrum over the region of the calculation is shown in the upper portion
of figure 3 and is indicated by Tj_.  We note that this sample is essen-
tially opaque (T, = 0) near the centers of the two absorption lines.
We assume^that the bandpass filter passes only the spectral region from
(v - va) = 0 to 4, which corresponds to the region between the centers of
the lines.

The attenuator on GC2 is a netural density filter with constant trans-
mittance, T2, equal to the average transmittance of the CO in correlation
cell GCl over the bandpass of the filter.  The third curve in the upper
panel of the figure is a plot of the difference between the other two
transmittance curves.  This curve represents the spectral response of
the system.  Since T2 is equal to the average of T1; the integral of
the curve (T2 - T^) is 0.  Thus, with no sample in the sample cell, no
voltage component Va is produced as the rotating cells move first GCl and
then GC2 into the beam.                  <

We now consider a sample cell containing enough CO to produce the trans-
mittance spectrum indicated by the center panel of figure 3.  This
corresponds to two of the stronger CO lines with an absorber thickness of
0.01 atm cm at a pressure of 1 atmosphere.  When GCl is in the beam, the
spectrum after the beam has passed through the sample is given by the
product of the transmittances TI x Tsam.  This is indicated by the curve
in the lower panel of figure 3 along with the corresponding curve that
applies when the beam is going through GC2.  The resulting detector sig-
nal, Va, is proportional to the difference between the two integrals
                                  -7-

-------
     1.00
    0.50 H-
w
o

    1.00
    0.50
               'Sam
Y
    1.00
    0.50
          v/  /
           I   /
           1   /
                          .'T, x T
                            2    sam
Figure 3
                   x T
                  1    sam
                                                         .
                                                         V
                        V-V   WAVENUMBER (cm"1)
                           a
           Calculated spectral plots of transmittance for a model
           spectrum similar to a portion of a CO band.  The wave-
           number scale is measured from the center of one of two
           identical lines.  The curves illustrate the spectral
           principle of operation of a gas analyzer using one gas
           correlation cell filled with the gas species being
           measured and an attenuator with a flat response whose
           absorptance is equal to the average absorptance of the
           gas correlation cell.

-------
J(T1 x Tsam) dv and ,f
-------
     1.00
    0.50
    1.00
H

g


I
    0.50
            T     =  T
             sam     X
    1.00f-
    0.50
   0



  -0.10
                  (T  -
x T
                               sam
                  -V-v  WAVENDMBER (cm"1)



  Figure 4.  Calculated spectral plots of transmittance

             demonstrating the response of the gas corre-

             lation system to a gas other than the one to

             be detected.
                     -10-

-------
in the sample.  For example, if I ,  the average transmittance of the
interferring sample is 0.8, the voltage Va produced by the CO is reduced
to 0.8 of its correct value.  It follows that changes in source brightness
or detector sensitivity would also change Va accordingly.

We now reconsider the system illustrated in figure 1 with the chopper
operating at a carrier frequency fc.  The detector signal, V ,  at
frequency fc is proportional to J*T2 x T8ain dv when GC2 is in°the beam
and is proportional to JT]_ x Tsa!L dv when GCl is in the beam.  The sum
of these two integrals, RTi + T2) Tsam dv is proportional to Vc, aver-
aged over a complete cycle of the pair of rotating cells.  Absorption
by an uncorrelated interfering gas reduces Vc by the same fraction that
it reduces Vai which is proportional to J(T2 - Ti)Tsam dv.  Thus the
ratio Va/Vc is unchanged by an interfering gas that is uncorrelated with
the CO, or by variations in source brightness or detector sensitivity.
The signal of primary importance is therefore Va/Vc, since it is a mea-
sure of the amount of CO gas and is insensitive to the variations that
produce serious errors in many systems.


An electronic circuit designed to measure Va/Vc is shown in figure 1.
When correlation exists between the absorption spectra of the sample
and the absorption spectra of the gas within the correlation cell, the
carrier is modulated at 30 Hz.  The phase of the modulation depends on
whether the correlation is positive or negative.  Since a synchronous
demodulator is used in the electronics, the presence of certain sample
gases may result in a negative signal output.

To simplify the measurement of Va/VC; an AGC (automatic gain control) is
employed within the electronics.  Both Va and Vc are amplified by the gain
controlled amplifier.  The gain of this amplifier is continuously adjusted
by the AGC to hold the average value of Vc constant.  The Va component of
Vc is directly proportional to the ratio Va/Vc when Vc is held constant.
The system is calibrated in terms of signal output with the AGC operating
versus the absorber thickness of the gas within the sample cell.  The
calibration is performed with a flowing gas sample at one atmosphere
pressure and at room temperature; the absorber thickness is proportional
to the gas concentration for a given sample cell length.
                                    -11-

-------
                              SECTION III

                                 OPTICS
 GENERAL DESCRIPTION

 The three-channel gas correlation spectrometer optics assembly was
 designed for good stability and large throughput while preserving a
 reasonably small package size and weight.  The selection of optical
 components was limited to off-the-shelf items wherever possible.  First
 surface mirror optics were used exclusively to eliminate chromatic
 aberation.  All imagery is performed with spherical mirrors.   All of
 the windows are sapphire and are antireflection coated to reduce reflec-
 tion losses and optical intereference between window surfaces.  Apertures
 reduce scattered energy; some are overfilled to increase stability.
 Components are arranged to minimize alignment errors that could result
 from flexure in the baseplate or from differential expansion  due to
 thermal gradients.   Except for the source,  all components that can be
 removed are either  self aligning or do not  require alignment.   All other
 components are bonded in place.   The system may either be used horizon-
 tally on a bench or mounted vertically within an electronics  rack.   The
 overall dimensions  of the basic  optics assembly as used on a  bench top
 are about 39 cm wide by 25 cm high by 100 cm long.  Angle brackets
 attached to the instrument sides with rubber loaded bolts allow the  optics
 assembly to be mounted in a standard 19-inch (48.3 cm)  electronics rack.

 A  schematic diagram of the optics  assembly  is  presented in figure  5.  The
 three-channel system has two major subdivisions;  the C02  system and  the
 CO and  Cfy systems.   Only the motor,  the  Nernst source,  the reference
 pickup,  and the bias  battery are shared by  these  two subsystems.   Each
 of these  subsystems  is further divided into the subassemblies  presented
 in figure  1 for a single channel system.

 SOURCE ASSEMBLY

A  single infrared source  is  shared by  all three channels  of the gas
correlation spectrometer.  The Nernst  glower source  has a relatively
high radiant emittance in all  three spectral regions.  It also has good
                                  -12-

-------
LO
I
                                                                                                         M23
            Figure 5.  Schematic diagram of the optics assembly.
                       The components are. numbered as a mirror M, window W;  thermoelectrically  cooled
                       photodetector D, filter F, plane diffraction grating  G, aperture A,  slit S,
                       gas correlation cell GC, or bearing block BB.

-------
 stability and long life, and is easy to use.  The standard source
 mount has been mounted on an assembly that allows transverse adjust-
 ment for proper alignment of the source.

 The source dissipates about 60 watts of power during normal operation.
 The source is surrounded by a double walled box.   A fan forces room air
 through the box between inner and outer walls to  keep the system cool.
 Gaskets prevent the cooling air from leaking into the optical path.   The
 cooling assembly is thermally isolated from the remainder of the optics
 unit.   The cooling assembly has been black anodized to increase its
 emissivity and absorptivity.   The wall that separates the source from
 the gas correlation cell assembly has been mirrored to reflect heat
 back into the source  cooling  assembly.   The power dissipation of the
 source is reduced  to  about  18 watts  when the source is operated in  the
 standby mode.  The fan speed  is reduced during this mode  of operation to
 prolong the fan  lifetime.

 GAS CORRELATION  CELL  ASSEMBLY

 An exploded view of a set of  rotating gas  correlation cells is presented
 in figure 2.   The  cells are pinned to an axial shaft that is driven  at
 1800 rpm by a synchronous motor.   This  has the effect of  switching  the
 optical path between  the cells  at 30 Hz.   Energy  from the source is
 chopped at 360 Hz  with an opaque 12-bladed chopper that is deposited  on
 the entrance window of the  rotating  cells.   The 360 Hz chopper is on  the
 source side of the gas correlation cells  to minimize scattered energy
 within the cells.   Cell GC2 contains a  non-absorbing gas,  N2,  and has a
 metallic coating of about 70% transmittance deposited on  one window.
 Cell GC1 contains  5%  N2 as well as the  two gas species being detected,
 32% CH4 and 63%  CO at a total pressure  of  0.71 atm.   The  concentrations
 and the total pressure were chosen so that the transmittances of the  CH4
 and the CO over  the corresponding spectral  band passes matched that of
 GC2.   The lengths  of  GC1  and  GC2,  1  cm, were selected so  that the total
 pressure of the  mixture in GC1  that  balanced the  transmittance of GC2
 was somewhat  less  than 1  atm.   The construction of the cells  is  simpli-
 fied by keeping  the internal  pressures  below atmospheric  pressure.  The
 windows  are pressed against the bond type  seal between the window and
 the cell  body by the  pressure gradient.  Optimum  performance  is  normally
 achieved  when the  gas  pressure  in GC1 is equal to or somewhat less than
 the pressure  of  the sample being  studied.   However,  the effect  of changing
 the total  pressure by  as much as  a factor  of 2 is slight  provided the con-
 centrations of the CH4  and CO are adjusted  to  balance the  transmittance
 of  GC2.

A similar pair of  correlation cells  is  used  for C0~.   Cell GC3  contains
0.66 atm of pure C02 and GC4  contains N2 at  approximately  the  same
pressure.  These cells  are 2  cm long.  One window of GC4  is coated with
an attenuator similar  to that on  the window  of GC2.   The gas  correlation
                                 14

-------
cells  are  contained within a dustproof chamber that is sealed with a
flexible foam  gasket  to prevent an accumulation of dust on the cell
windows.

A  single-bladded  chopper pinned to the correlation cell shaft interrupts
a  beam of  light between a light emitting diode and a phototransistor.
The 30 Hz  output  of the phototransistor is used as a reference source to
allow  synchronous demodulation of the 30 Hz detector signals.

The motor  shaft is connected to the gas correlation cell shaft with a
flexible coupling.  The motor and the bearing blocks BB are pinned to
the baseplate  to maintain proper alignment.  The motor is mounted within
a  chamber  in the  optics assembly.  Room air is drawn through the baser
plate  and  through the motor by a fan in front of the armature and ex-  '
hausted back through  the baseplate.  This airflow removes the heat dissi-
pated  by the motor without circulating room air through the optical path.
The amount of  heat dissipated by the motor is reduced by operating it
at 80  V ac.  This reduction in operating voltage also reduces the amount
of vibration generated by the synchronous motor without changing its
speed.

C02 SYSTEM

The source is  Imaged by spherical mirror M51 through a set of gas corre-
lation cells onto aperture A54, which is mounted on window W51.  The
source energy  transmitted by A54 is imaged by spherical mirror M55 through
the sample cell and onto the plane of window W53.  This energy is collected
by spherical mirror M58 and directed onto photodetector D51 through band-
pass filter F51.  The passband of this interference filter is centered at
about  4.405nm  (2270 cm'1).  The width of the passband between the 1/2
maximum transmittance points is about 0.044nm (22 cm"1).  The passband
includes most  of  the P-branch of the V3 absorption band of the 13C ^02
isotope.  Photodetector D51 includes a PbSe element mounted on a single
stage  thermoelectric cooler.

Aperture A51,  which is mounted on the face of M51, is imaged by mirror
M55 onto the plane of window W52.  Aperture sizes, focal lengths, and
dimensions were adjusted such that the image of A51 on W52 is the same
size as the image of A54 on W53.  These images are smaller than the
inside diameter of the sample cell.  The 4 cm long sample cell  has
the same inside diameter as the 20 cm cell.  The image of A51 on W52 is
re-imaged by M58 onto the photodetector.

Both window Wl and mirror M52 are mounted in the wall of the correlation
cell chamber near the source assembly.  This wall has also been relieved
to allow for thermal expansion without affecting the adjustment of M52.
Further, M52 is shielded from direct source radiation by a wall of the
source assembly.  Apertures A52 and A53 minimize scattered energy, and
                                 -15-

-------
   aperture A54 is overfilled to increase the stability.  The limitine
   apertures of this  system are A51  and A54.                  A unit ing
   It  is  expected  that  the  4 cm long sample cell will be used for samples
    sr»
                     s:
                                                             -
  cell.  The flow is directed through the shorter cell first to allow
  observance of abrupt changes in gas concentration before mixing can
  dowT,   ^ 10n8^ Cel1'  The S3S fl°W is directed near tne"ell  win-
  dows to prevent volumes of stagnant gas at the ends of the sample  cells.
  CO and CH^ SYSTEMS
General

The
the
             ?4 detection systems are discussed together since most of
               P      nd °tlCa
  te c™
  T   ZP^f   P      fnd  °PtlCal  ray P3ths are C0iranon  to b°th systems.
  The specific  spectral passbands  employed are controlled by the grating
  fSS  aST   y  "u C0rabination wi<* an interference filter near each
  photodetector.   These passbands were selected and adjusted to give a
  low8* thj°?**ut' 8°od sensitivity to the gas being detected, S a very
                                                   *
                                                             ,
                    °ther 8aSeS that mi§ht be in th  sample gas.  The
                                   ered at about 3.315^, (3017 cm-1)
                                    the 1/2 ™*^ tra^ittance points
 Q-branch of   *              ,    passband is almost centered on the
 Q branch of the 3.3[m absorption band of Cfy.  The passband of the CO
 S                                   (2171 «-1)*  w^h a wJdth of about
               -
 l2 oTh  f ^      ?   PaSSband lncludeS abs°rption lines Rl through
R12 of the fundamental absorption band of CO.                    "«ugn

As in the C02 system, two sample cell lengths are employed to extend the
useful range of sensitivities of the CO and CH4 systems.  Both smple
   iS.TuaCtiVe 3t aU times'  The existinS s^ of electronics Sn be
switched between the outputs of the photodetector preamplifiers wi?h
switches on the electronics units.   With this optical syste^ It would

          6  °d  ^
                                                           e     wou
 coTo!^6   °ifd?  ^  ^lectr°nic  a^omatic range  switching devidto
 control  the  cell  length  in use.   Alternately,  the outputs of the photo-
 detectors used with both cell lengths can be'monitored s imul taneoSs 1 y
 for flowing  gas samples  which have rapid and very large fluctuations in
 the gas  concentration.   About 75  percent of the radiant energy trans-
mitted by the grating- filter assembly is directed through thf 100 «
sample cell.  The remaining 25 percent is directed through the 10 cm
sample cell.  This division of energy is based on the sifnal-to-noise
requirements of the system.                                      uu^e
                                  -16-

-------
Source Optics and Grating Filter Assembly

The spectral passbands of both the CO and the CH,  systems are set by the
grating-filter assembly.  This grating monochromater forms an undispersed
image of the entrance slit, Si, on the exit slit,  S4,  of the unit.

The source is imaged by mirror Ml through a pair of gas correlation cells
onto slit SI.  The energy transmitted by SI is collimated by spherical
mirror Mil and directed to the plane diffraction grating G.  The grating
is overfilled and acts as a limiting aperture.  Part of the energy that
is dispersed by  the grating is collected by spherical mirrors M12 and M13.
These mirrors form many dispersed monochromatic images of SI.  Radiant
energy near 4.6 pm is focused onto slit assembly S2 by mirror M12.  Simi-
larly, energy near 3 |_un is focused onto slit assembly S3 by mirror M13.


 Figure  6  shows  an expanded view of  slit  assemblies  S2  and S3.   The
 energy  is  reflected  vertically from the  bottom mirror  through  the slit,
 and  reflected  from the  top mirror back to  the  corresponding  mirror M12
 or M13.   The dispersed  energy that  is  transmitted by  S2  and  S3 is
 de-dispersed upon the second  pass through  the  grating  G.   Mirror Mil
 forms an  undispersed image of the entrance slit SI  onto  the  exit slit  S4
 through flat mirror  M14.   Slits SI,  S2 and S3,  and  S4  are conjugate to
 each other.  The spectral widths of  the  two passbands  are set  by the
 widths  of  S2 and S3.  The center wavelengths of the passbands  are set
 by the  angular  locations  of S2  and  S3.   The slopes  of  each side of  the
 passbands  are  set by the  widths  of  the images  of either  SI or  S4 on S2
 and  S3.   Slit  SI was made slightly wider than  S4 to simplify alignment,
 thus  the  slopes  are  set by the  image  of  S4,  which  lies within  the bounds
 of SI.

 The  central ray  of the  energy transmitted  by S4 is  below the central
 ray  transmitted  by SI.  These rays  are parallel to  each  other  and to
 the  optics assembly  baseplate.   The  central rays converge from Mil  to
 the  center of  the grating where  they  cross.   They are  divergent from
 the  grating to slit  assemblies  S2 and  S3.   The limiting  apertures of  the
 grating-filter assembly are slits S2,  S3,  S4,  and the  grating  G. Aper-
 tures Al, All, A12,  and A13 reduce  the scattered energy.

 The  grating-filter assembly transmitts undispersed  radiant energy at
multiple  orders  of the  desired  spectral  passbands.  Interference filters
 are  used  to  limit each  passband  to  a  single order of  the grating.  A
 pair  of these narrow bandpass filters  are  located near the photodetectors
 of each channel.   This  restricts the  amount of ambient radiation reaching
 the  photodetectors.  One  of these filters  is also used as a  dichroic
mirror to  separate the 4.6 ^m radiant  energy from the  3  u.m radiant  energy.

 The components in the grating-filter assembly are enclosed within a
 dust-proof chamber.   The base of this chamber serves as  the mounting
                                 -17-

-------
oo
                                                                                       MIRROR
                                                                                       SLIT
                                                                                      MIRROR
                     Figure 6.  Expanded view of slit assembly S2 or S3.  Many monochromatic
                                images of the entrance slit SI are formed on slits S2 and S3.

-------
platform for  these components.  This chamber is, in turn, mounted on
three  spring  loaded points against the optics assembly baseplate.  A
ball and spring-loaded-post maintain correct alignment between the
grating-filter assembly and the remainder of the optics assembly.  This
mounting technique isolates the grating-filter assembly from stresses
that may be applied to the optics assembly baseplate.  This isolation
assures  stable wavelength passbands for the CO and the CH^ systems.

Sample Cell and Photodetector Assemblies

The optical path  length between the short sample cell entrance and
exit windows W21  and W32 is 10.0 cm.  The radiation that exits through
the lower 25 percent of slit S4 is intercepted by flat mirror M31 and
directed to spherical mirror M33.  An image of mirror Mil is formed on
the entrance window W21 of the sample cell.  An image of slit S4 is
formed on photodetectors D31  and D32.  Interference filter F31 reflects
3 urn energy and transmits radiant energy near 4.6 pm.  Interference
filters  F31 and F32 are bandpass filters that select a single order of
the energy transmitted by the grating-filter assembly.  They also reduce
the amount of scattered energy reaching the photodetectors.

The optical path  length of the double-passed sample cell is 100.0 cm
The energy that exits through the upper 75 percent of slit S4 is collected
by spherical mirror M21.  The top edge of mirror M31 defines the lower
boundary of slit  S4 as viewed from mirror M21.  Both M31 and S4 are mounted
on window assembly W12 to maintain stable slit dimensions.  An image of
mirror Mil is formed on window W21 by spherical mirror M21.  Similarly,
an image of S4 is formed on mirror M23 by mirror M21.  Spherical mirror
M23 forms  another image of Mil beside the image of Mil on W21.  The
image  of S4 on M23 is re-imaged on photodetectors D21 and D22 by spheri-
cal mirrors M24 and M25, respectively.  Interference filters F21 and
F22 serve  the same functions as filters F31 and F32.  Photodetectors
D21, D22,  D31, and D32 include PbSe elements which are mounted on two-
stage  thermoelectric coolers.

The sample cell windows  and end mirrors are easily removed for cleaning.
The windows are bonded to flanges.  The end mirror is covered by a
metal  cap.  Rubber "0" ring seals with metal-to-metal contact between
the cell and the  end cap or window flanges provide a gas tight unit.
Similar  construction  is used for the windows on the C(>2 sample cells.
Mirror M23 is self-aligned by spring loading it against three points
that contact the  front spherical surface.   The sample cell temperature
may differ from the baseplate temperature as a result of the sample
temperature.  The sample cell  is held rigidly to the optics baseplate
near W21.  The other end of the cell is allowed to mo/e along the cell
axis.  This allows for stress-free differential thermal expansion of
the cell.  Mirror M23 is mounted to the sample cell.  Since the cell is
mounted  to the baseplate near W21, which is near the center of curva-
ture of M23, the alignment is insensitive to small movements of the
sample cell relative to the baseplate.

                                  -19-

-------
Apertures  A21,  A22,  and A23  reduce  the  amount  of  scattered  energy  that
can reach  the  photodetectors.   In addition, A21 protects  the surfaces of
M21 and M22  during  the removal  and  reinstallation of window W21.

The sample gas  coming  from the  C02  cells flows into and through the
10  cm  cell and  then  into the 100  cm cell.  It  is  exhausted  from the
system after flowing lengthwise through the 100 cm cell.  The sample is
directed through the shorter cell first to allow  the observance of
abrupt changes  in gas  concentration before any mixing can occur in the
longer cell.  Although not shown  in the figure, a small portion of the
sample gas flows  into the cell near W21.  This prevents the possible
formation  of a stagnant pocket  or eddy  of gas near W21.  Similarly,
although most of the sample  is  exhausted from the cell just in front
of M23, a small flow of gas  is  exhausted from behind M23.   This prevents
the small volume between M23 and the cell end cap from containing
stagnant gas that could contaminate samples added later.
                                 -20-

-------
                               SECTION IV

                              ELECTRONICS
GENERAL DESCRIPTION

The electronics for the three-channel gas correlation spectrometer were
designed for good stability and low noise while preserving a reasonably
small package size and weight.  The photodetector is the limiting noise
source in each channel.  Component selection was limited to off-the-
shelf items.  The system was designed to operate within a laboratory at
room temperature.  To minimize costs, the electronics for each of the
three channels are as much alike as possible.

A  block diagram  of the  electronics for a single  channel is presented in
 figure  1.   The photodetector  output  consists of  a 360 Hz carrier modula-
 ted at 30 Hz.  The ratio of the modulation amplitude to the  carrier
 amplitude is directly proportional to the amount of gas within the
 sample cell.  Phase  locked  synchronous demodulation of  the 30 Hz  signal
 offers  the  best  system  signal-to-noise.  The 30  Hz reference required
 for synchronous  demodulation  of  the  30 Hz signal is also employed to
 synchronously modulate  the  carrier as necessary  to cancel  any residual
 30 Hz modulation on  the carrier  when no  sample is present.

 The three electronics units and  the  power supply unit are mounted in a
 standard  19-inch (48.3  cm)  electronics rack.   The front panel heights
 are 8.75  inch  (22.2  cm) for each electronics unit and 7 inches  (17.8 cm)
 for the power  supply unit.  Line power is furnished  through  a single
 junction  box within  the electronics  rack to  prevent  possible ground
 loops between  systems.   The power supply unit  is mounted  in  the  top of
 the electronics  rack to prevent  the  heat dissipated  from  this unit from
 unnecessarily  heating the  other  units.   The  front  panels  of  the  power
 supply  unit and  one  of the  electronics  units are presented in figure 7.
 The  signal  and power cables that connect the optics  assembly to  the
 electronics unit are about  3.5 meters  long.
                                  -21-

-------
                                   '
                                 sim *•'«"'
         PHILCO
                                                     /
                                                                0
  —— GAIN 	   	'•'   I— MODULATION
  SIG I    SIQS   HBF   SIGNAL  CASHIER
Figure 7.   Power supply unit above a single electronics  unit.
                                -22-

-------
OPTICS UNIT

A schematic illustration of the electronics contained within the optics
assembly that are used by a single signal channel of the system is pre-
sented in figure 8.  The gain of each preamplifier is set to effect a
carrier signal amplitude that is about 1 V peak-to-peak.  Five photo -
detectors and preamplifiers are used for the three channel system.

A 45 V dc battery  pack supplies the bias voltage to  the photoconductive
PbSe detectors.  Jacks allow monitoring of the battery voltage in both
the load and the no-load conditions.  The total current drain from the
bias battery by the photodetectors is about 40 u-A when the photodetec-
tors are thermoelectrically cooled.  This should allow a battery  life
in excess of six months.  A low-pass filter with a 1.1 second time con-
stant in each channel prevents a possible crosstalk  between channels
as a result of coupling through the bias voltage supply.  This filter
also reduces any noise generated by the bias supply.  A diode in  series
with the battery protects the electrolytic filter capacitors from
damage that could  occur if the battery polarity is reversed.  A
capacitor across the diode lowers the impedance of the protection cir-
cuit.  The batteries are contained within a fluid tight chamber to
prevent possible contamination of the optics system  with battery
electroyte.

The 30 Hz reference pickup shown  in figure  1 consists of a  light
emitting diode  (LED) and a phototransistor used in  the  common emitter
mode.  The beam of light between  the LED and the phototransistor  is
chopped at 30 Hz by a single-blade chopper mounted  on the correlation-
cell shaft.  The output impedance is about 490 .n. and the amplitude of
the reference signal is about 400 mv peak-to-peak.

The reference system is supplied with  -15 V dc  from a power supply in  an
electronics  unit.   The photodetector preamplifiers  are supplied with +15
and  -15 V dc from  the same power  supply.   Separate  cables connect the
reference and the  preamplifier  systems  to  the  electronic unit  power
supply.  Further  isolation between  systems  is  accomplished  with resis-
tance-capacitance  filters  that  are  installed next  to the reference
pickup  and  the  preamplifiers.

POWER SUPPLY UNIT

The Nernst  source, source  heater,  and  the  correlation cell  drive motor
receive  their  60 Hz power  from the  power supply unit.  The  source
current  is  controlled  at  the power  supply unit and is displayed on a
 front panel meter.  The  amount of current  that flows through the  source
is  limited  by  either one,  two,  or three ballast tubes that  are switched
                                  -23-

-------
  TO OTHER
  CHANNELS
                                      DETECTOR
                                      BIAS
                                                       NEDA  #709
                                                       45  VDC  BATTERIES
                         BATTERY
                         TEST
                         JACKS
                                                                TO  ELECTRONICS
                                                                          UNIT
PHOTODETECTOR ON   '
THERMOELECTRIC
COOLER             I 	
PREAMPLIFIER    |
       Figure 8.  Schematic illustration of a single channel of the
                  photodetector electronics within the optics assembly.
                                     -24-

-------
 in parallel between the source and the 115 V ac power line.   The fan
 that cools the Nernst source is controlled by this same switch.   The
 fan is operated at 115 V ac except in the standby (STBY) mode when its
 voltage is reduced to about 70 volts.  This approximately halves the
 fan speed.  The source heater is supplied from a 60 V ac transformer in
 the power supply unit when the spring loaded source start switch is  held
 on.  In normal operation, the synchronous 1800 rpm motor is  powered  by
 80 V ac supplied by a transformer in the power supply.   The  motor voltage
 is 115 V ac when the spring-loaded motor switch is held in the start
 position.  The thermoelectric coolers for the photodetectors are powered
 from a 5 V dc power supply mounted in the power supply  unit.  A  3 n
 power resistor in series with each thermoelectric cooler converts the
 fixed-voltage supply to a fixed-current  supply.  Each of the five coolers
 draws about 1.8 amperes.

 ELECTRONICS UNIT

 General

 Each of the three gas channels employs an electronics unit to process
 the detector signals.  A functional block diagram of a  single unit is
 presented in figure 1.   The electronic unit is subdivided into a 30  Hz
 reference section,  360 Hz carrier section,  and a 30 Hz  signal section
 Most of the electronics for these sections  are contained on  plug-in
 circuit cards  with  one  card for each section.   The  circuit cards  are
 interchangeable  between the three electronics  units.  The differences
 among the three  units are in the front panel  controls.   These differences
 reflect the use  of  a  single photodetector for  the  C02 system and  a pair
 of  photodetectors for the CH4  and CO systems.   Switching between  photo-
 detector  preamplifiers  is accomplished with a  toggle  switch  on the front
 panel of  the electronics unit.   Figure 7  illustrates  the front panel
 controls  for a unit for use with either the CO or  the CH4 systems.

 Reference  Section

 The  reference  section of the unit is  used to amplify, phase  shift, and
 transform the  signal  from the  reference pick-up into  a  form  compatible
with  the  remainder  of the electronics  unit.  The input  reference  signals
 are amplified  to a  set  level by the  first stage  of  the  reference  section.
The gain  of this amplifier  is  adjusted by the  operator  and is  not  critical.
The reference  signals are  then  further amplified by a tuned  amplifier
with  a  Q of 5.  This  amplifier  is  followed  by  two adjustable  phase shifters
connected  in series.  The  second  phase shifter  thus tracks any adjustment
made  in the first one.   The  phase  shifters  are  used to  obtain  the  proper
phase between  the reference  and  the  photodetector signal.  The phase can
be adjusted easily without  test  equipment and  should  remain  stable.  The
output stages  of the phase  shifters  transform  the sinusoidal  reference
signals into large amplitude square wave  reference  signals that drive
the gates of field effect transistor  (FET)  switches in  the modulation
balance assembly and in  the  30 Hz  synchronous demodulator.


                                 -25-

-------
 Carrier Section

 The 360 Hz carrier signal section amplifies and normalizes the signals
 to a constant average carrier level.  This section is also employed to
 balance out erroneous signals that can result from an imbalance in the
 transmissivities of the correlation cells.

 The input signal from the photodetector preamplifier is amplified with
 a bandpass amplifier.  The Q is  about 3.5 and the center frequency is
 360 Hz.  The operator sets the gain control potentiometer of  this ampli-
 fier to effect an indicated signal level that is near midscale on the
 front  panel meter when the sample cell is filled with clean air.   The
 adjustment is  not critical since the automatic gain control (AGC) has  a
 dynamic range  of more than 5 to  1.   The gain controlled stage of  the AGC
 follows the tuned amplifier.

 The modulation balance assembly  follows  the  gain controlled amplifier.
 This assembly  consists of  an adjustable  voltage  divider and an FET
 switch  that  shunts part of the carrier  signal to ground when  the  switch
 is closed.   The  FET switch is activated  at 30 Hz by  the reference sys-
 tem.  This  results in  a square wave modulation at  30  Hz of  the 360 Hz
 carrier.  The  voltage  divider is  adjustable  to allow  the  FET  to shunt
 up to 10 percent  of the carrier  signal to ground when  the switch  is closed.
 The adjustable system  includes two potentiometers  mounted on  coaxial
 shafts.  One potentiometer is for fine adjustments, the  other  for coarse
 adjustments.   The adjustment resolution  of the system  is +2 x  10"5 of the
 carrier amplitude.  A  front  panel mounted switch allows  the modulation
 balance system to be turned  on and off.  This  three position  switch also
 allows  the  inserted modulation to be either in phase or  180 degrees out
 of phase with  the 30 Hz signal modulation.

 The 360 Hz  carrier demodulator follows the modulation balance  assembly.
 The demodulation  is performed by a precision full-wave  rectification
 circuit.

 The average demodulator output voltage is compared with  a fixed ref-
 erence  voltage by the  AGC  control amplifier.  Any  error between the two
 voltages is coupled back to  the  gain controlled  amplifier with a  feed-
 back loop.  The closed  loop  response time constant is one second.
 Shorter time constants would remove an excessive amount of  the 30 Hz
 signal modulation from the carrier.  Longer time constants would
 increase the fractional error in the 30 Hz signal modulation that
 could occur due to large abrupt changes in transmissivity of the sample
 gas.  The output of the carrier demodulator can be monitored on the
 panel meter and can be recorded from the carrier output jack.   The AGC
can be turned off with a switch on the front panel.  The gain of the
gain controlled amplifier is then set by a potentiometer that  is mounted
                                 -26-

-------
beside the switch.  With the AGC off, the 360 Hz system output is  a
measure of the sample gas transmissivity over the spectral bandpass of
the system.

When the manual gain control is set to effect the same average carrier
level as is maintained by the AGC, then the sensitivity of the 30  Hz
system to small concentrations of sample gas is the same as when the
AGC is operating.  With this gain setting, the operator can block  the
radiation between the source and the photodetector to observe the
character and the relative amplitude of the noise generated by the
photodetector and the electronics systems.

Signal Section

The 30 Hz section of the unit amplifies and isolates the 30 Hz signal
modulation.  This is performed with a phase locked synchronous ampli-
fier and demodulator tuned to 30 Hz.  The demodulated output of the
carrier serves as the input to the 30 Hz signal section.  Much of the
720 Hz ripple that results from the 360 Hz demodulation process is
removed by a three-stage low-pass filter.

The range switch  is a stepped attenuator that sets the signal output
range.  The signal gains of the attenuator are XI.0, X0.3, XO.l,
•  • •, X0.0001,  and off.  The off position is used in setting the  zero
output level of  the 30 Hz system.  A single curve of the sensitivity
of the system to  different amounts of sample gas can be drawn by
normalizing the  output signal Va/Vc by the attenuator gain g.  The
calibration potentiometer is a variable gain control that follows
the range  switch.  A calibration  constant Ca is proportional  to this
gain.  It  is adjusted by the operator such that the system response
curve of V* =  (Va/Vc)/(g Ca) versus  gas concentration is  in  agreement
with the  indicated output for the calibration  sample  gas  that is
employed.

The calibration  potentiometer  is  incorporated  into the  feedback  loop
of a high gain bandpass  amplifier with  a  Q  of  8.   The signal is further
amplified and conditioned before  being  demodulated.   A  notch filter in
front of  the demodulator reduces  any 60 Hz modulation that might  be
present  on the carrier.  The  synchronous  demodulator  only demodulates
the 30 Hz portion of  the signal  that is  in  phase with the 30 Hz
reference.  Any  residual 60 Hz  signal is  passed by the  synchronous,
phase-locked, demodulator as  ripple  and does  not affect the  smoothed
signal  output.   An  amplifier  with an adjustable time  constant smooths
the demodulated  signal  before it  is  presented on the  front panel  meter
or at  the signal output  jacks.   The  time  constants available are  0, 0.1,
0.3,  1.0,  and  3.0 seconds.   In the  0 position the response time is
 limited  to about 0.08 second  by the  combination of low-pass  filters,
notch filters, and bandpass  amplifiers.
                                 -27-

-------
A meter mounted on the front panel of the electronics unit allows the
operator to monitor the reference level, the input signal level, the
carrier level, and the output signal level.  The reference and input
signal levels are adjusted to effect a mid-scale indication on the
meter.

Identically labeled input and output jacks are provided on both the front
and back of the electronics units.  The jacks are wired in parallel.  A
10 K n resistor is in series with each pair of output jacks.  The outputs
can be grounded without affecting the operation of the system.  The full
scale signal output is -1.0 volt.  The normal carrier output is about
-1.5 volts.  Shunt resistors across the output jacks will reduce these
voltages to match the ranges of low level recorders.  For example, a 10 Q
shunt resistor will act as a voltage divider with the 10 K fi resistor to
reduce the full scale signal output from -1 volt to -1 millivolt.

For clarity, the above description was limited to a single channel of
electronics.  Since two different sample cell lengths are employed for
each gas species, two separate modulation balance assemblies are
included in each electronics unit.  The CO and the CHA detection systems
also employ two different photodetectors and preamplifiers for each gas
species.  Separate adjustable gain controls were provided for each
signal input.  Toggle switches are used to switch from the signal inputs
and modulation balance assemblies used with the short sample cell to
the corresponding ones for the long sample cells.  This construction
allows the operator to switch rapidly between the different sample cells
without changing the settings of the signal input gains or the modulation
balance assemblies.
                                 -28-

-------
                              SECTION V

                              TEST DATA
The sensitivity of the gas correlation spectrometer to various  gases
has been measured.  Plots of the normalized output, y* , versus  the
sample concentration are presented in figures 9,  10,  and 11 for CO, CH4,
and C02> respectively.  For a given gas concentration, the system out-
put is a function of the setting of the calibration potentiometer and
the range switch.  Both of these controls adjust the gain of the 30 Hz
system.  The calibration controls were set to effect a full scale out-
put of about -1 volt for sample gas concentrations of 100 ppm of CO,
50 ppm of CH^, and 1% of 0)2 when the samples were in the long sample
cells.  With the AGC operating, the output of the system Va/Vc was
then normalized by the gain setting, g, of the range switch used for
each measurement and plotted against the concentration of the sample
gases.  The ordinate of the curve V* is equal to (Va/Vc)/(g Ca) , where
Ca is a calibration constant that is proportional to the gain of the
calibration control.
The sample gases were prepared by diluting measured amounts of CO,
and C02 with N2«  Some of the samples contained both CO and City with
the N2 dilutant.  The sample containers were evacuated before use to
prevent contamination of the sample.  The samples were mixed with a
mechanical stirring device inside the sample tank.  The sample gas flow
rate through sample cells was regulated at 0.2 liters/minute while V*
was being recorded.  This flow rate is fast enough to prevent adsorption
or desorption of gas from the walls of the sample cell and the gas
handling system from affecting the concentration of the sample gas.  A
slow flow rate also minimizes the amount of sample gas required for each
data point.  A much faster flow rate was used to flush an old sample
out of the sample cell with a new sample to reduce the waiting time
between measurements.  The gas flow rate is limited by a needle valve in
the input gas line.  The flow exhausts from the sample cell to the atmos-
phere, thus the pressure within the sample cell is slightly over one
atmosphere.
                                 -29-

-------
         100
o
                                                                           I  M HW- -I-I J_IJJ_U
                          100 CM SAMPLE CELL
                                                         10 CM SAMPLE CELL
            100            1000
    CO SAMPLE CONCENTRATION
Figure 9.  GO detection sensitivity.
                                                                                      ppm

-------
 1000 p	1  I  I I Mill	1   |  | I llll[	1   |  I I Hill	1   I I  I Mill	1—| II MM.
  100
*>
S
NJ
  0.1
0.01
                 100 CM SAMPLE  CELL
                                              0 CM SAMPLE CELL
I n. i     I   I  I Min _ |   |  | | |||| - 1   I I  I Mil - 1   I  I I III
                                                            ill
   0.1
                 10
                                              100
1000   ppm    104
                            CH4  SAMPLE CONCENTRATION



                     Figure 10.   CH^ detection sensitivity.
                                     -31-

-------
1000
               20  CM SAMPLE CELL
                                                   I  I I Mill	l'l|
                             0.1
                            1000
 1
104
10    7.       100
105  ppm      106
                         COo SAMPLE CONCENTRATION
                   Figure 11.  CQ* detection sensitivity.
                                -32-

-------
Two sensitivity curves are shown for each gas,  one curve for  each cell
length.  Both curves have the same shape but are displaced along  the
abcissa by the ratio of the cell lengths.  The  displacement results from
the output being directly proportional to the product of the  gas  con-
centration and the sample cell length.  This product is the absorber
thickness at a given temperature and pressure.   The useful measurement
range is shown by the solid line portion of each curve.  The  lowest con-
centration indicated is the noise-equivalent-concentration (NEC)  of the
channel.  The NEC is defined as the concentration that produces a signal
equal to the peak-to-peak noise level.  The upper limit of each measure-
ment range is set by the concentration where the slope of the logrithmic
curve is about 0.5.  This correspondends to concentrations for which v*
is proportional to the square root of the concentration.  Larger  gas con-
centrations can be measured, but the accuracy decreases.

The ability of this system to measure small amounts of one gas in the
presence of large amounts of other gas species  is one of the  main
advantages of the gas correlation system.  With the exception of  water
vapor, the system is insensitive to other gases in automotive exhausts.
For typical exhaust gas concentrations, the error produced by the inter-
fering gases was either less than the NEC or less than 1% of  the  amount
of gas being measured.  The discrimination ratio is given by  the  ratio
of the amount of interfering gas to the error signal produced by  the
interfering gas.  The system exhibits a discrimination ratio  of 50,000
against H^O by the CO system.  This means that it takes a sample  con-
taining 50,000 parts of 1^0 to produce the same signal as a sample
containing 1 part of CO.  The CE^ system has a discrimination ratio of
13,000 against H20.  A discrimination ratio with a negative sign means
that the interfering gas produces a signal of opposite polarity to the
signals produced by the gas species normally being detected.   Since the
error signal may not be directly proportional to the concentration of
the interfering gas, discrimination ratios are measured with concentra-
tions that are similar to those expected in automotive exhaust samples.
Large concentrations of interfering gas may produce an unrealistically
large discrimination ratio.

The discrimination ratios for several individual hydrocarbon species
were measured by placing a pure sample of the interfering gas in a 0.1 cm
long sample cell.  This flow-through sample cell was  located near the
exit slit of the grating-filter assembly in the beam of radiation that
traverses the 100 cm sample cell.  The 100 cm sample cell was filled with
N£«  A sample with a concentration of 1000 ppm in the  100 cm cell will
produce the same output as the pure sample in the 0.1  cm cell.  The use
of pure samples simplifies the gas handling procedure.  The hydrocarbon
discrimination ratio is based on the required number of carbon atoms in
the interfering gas to produce the same  signal as a single carbon atom
in a methane sample.
                                  -33-

-------
 A hydrocarbon mixture was prepared by vaporizing 31.2 microliters of
 Indolene Clear 91 octane gasoline into a 47.3 liter tank containing N0
 at J pressure of 1.51 atmospheres.  This lead free gasoline was analyled
 by EPA^on 11 May 1973.  At one atmosphere pressure, the mixed sample con-
 tains 486 ppmc in N,,.  This mixture in the 100 cm sample cell of the CH,
 analyzer produced an indicated output of -0.4 ppm of CIU.  This corres-
 ponds to a discrimination ratio of -1200.  Adding 50 ppmc of ethylene and
 50 ppmc of acetylene to this Indolene Clear mixture results in a hydrocarbon
 mixture that is representative of typical automotive exhaust minus the
 methane content.   Increasing the carbon content of  the Indolene Clear
 sample from 486 ppmc to 586 ppmc by the addition of ethylene and acetylene
 to the dilutant gas  does not change the indicated output of the system
 Thus,  the discrimination ratio against typical exhaust minus methane is
 about -1450.   This hydrocarbon mixture did  not produce a detectable  out-
 put  on the CC-2  or  the CO channels of  the system.

 Table  1 presents the  noise  levels  of the  system.  Each  channel  is
 detector noise  limited.  The peak-to-peak noise for a  1  second  time
 constant is expressed in terms of  the NEC.  The noise  levels  are approxi-
 mately inversely proportional  to  the square root of the  time  contants.
 The NEC values for the short sample cells are  larger than the NEC values
 for the long sample cells because the absorber thickness for a given gas
 concentration is less by the ratio of the cell lengths.  Further, the
 optical throughput of the 10 cm sample cell is only about 1/3 of the
 throughput for the 100 cm sample cell and the photodetectors with highest
 detectivity were used with the long cells.
 evewM               exhibits a slow drift in the zero signal output
control ad tr  ,* P^iodically corrected by the modulation balance
control adjustment.  The drift rate is rapid during the first few
Adrift  °rrfi0n after tUrninS on bot* ^e source and the motor.
The drift rate decreases to a very small value after about 2 hours
of warm-up operation.   The zero-drift was less than the equivalent of
1 ppm of gas in the 100 cm sample cell for the CO and the CH/, systems
over a 2-1/2 hour period after a warm-up period of equal time.   The
average drift rate after this period was thus less than 0.4 ppm/hour.
                                 -34-

-------
                                         Table  1.   SYSTEM NOISE LEVELS
i
10
Gas
CO
CO
CH,
CH4
Cell Length
(cm)
100
10
100
10
Time
Constant
(sec)
1
1
1
1
NEC (Noise-Equivalent
Concentration)
(peak-to-peak)
0.5 ppm
35 ppm
0.3 ppm
30 ppm
                                                                                       Norma 1

                                                                                   Operating Range
1 to 1,000 ppm




100 to 10,000 ppm




0.5 to 1,000 ppm




100 to 10,000 ppm
                             20               1           10 ppm (0.001%)           0.002 to 5%




                             4                l           50 ppm (0.005%)           0.01 to 25%

-------
                             SECTION VI




                        GLOSSARY OF SYMBOLS






  A# - an aperture




  ac - alternating current




 AGC - automatic gain control




  BB - a bearing block




  Cfl " calibration constant proportional to gain of calibration control



  cm - centimeter




  (im - micrometer




cm   - reciprocal centimeter,  a wavenumber unit



 CH^ - methane




  CO - carbon monoxide




 C02 - carbon dioxide




  D# - a  photodetector




  dc - direct current




  F# -  a  filter




  f   -  30 Hz  correlation cell  frequency




  f   -  360 Hz carrier frequency




FET -  field  effect transistor
                                -36-

-------
   g -  relative gain of 30  Hz  signal  attenuator



   G -  a plane  diffraction  grating



 GCl -  gas  correlation  cell containing  CO and



 GC2 -  gas  correlation  cell containing  N2 and a neutral attenuator



 GC3 -  gas  correlation  cell containing  COo



 GC^ -  gas  correlation  cell containing  N2 and a neutral attenuator



  Hz -  Hertz



  Kn •  kilo ohms



   0 -  ohms



 LED -  light emitting diode



  M# -  a mirror



  mV -  millivolt



  N, -  nitrogen



 NEC -  noise equivalent-concentration:   the gas  concentration that  will

       produce the same output signal as the peak-to-peak noise level



FbSe -  lead selenide



 ppm -  parts per million



ppmc -  parts per million of carbon



   Q -  measure of bandpass  of a tuned circuit



 rpro -  revolutions per minute



  S# -  a slit



  Tj_ -  transmittance of GCl



  T2 -  transmittance of GC2



 T   -  T    when sample gas is CO
  co    sam


T    - transmittance of sample gas
 sam
                                  -37-

-------
    TY " Tsam when samPle is s°me interfering gas



     V - Volt




    V  - normalized output (Va/Vc)/(g C )




    Va - 30 Hz signal amplitude




    Vc - average  360 Hz  carrier  amplitude




    W# - a  window




     % -  percent




    V  -  wavenumber




   va  -  wavenumber at the center of an absorption line




   Av  -  bandpass expressed in wavenumbers







                    ABBREVIATED LABELS ON CONTROLS






  AGC - automatic gain control




  CAL - calibrate




FUNCT - function




  MAN - manual gain,  the AGC is  turned off




  REF - reference




  SIG - signal




 STBY - standby




 T.C.  - time constant




   0  -  phase
                                -38-

-------
                               SECTION VII
                            PATENT INFORMATION
Follwoing are two (2) inventions which have been declared  as  subject
inventions:
  Name of Inventor

Dr. Darrell E. Burch,
Francis J. Gates,
David A. Gryvnak, and
John D. Pembrook

Dr. Darrell E. Burch,
Francis J. Gates,
David A. Gryvnak, and
John D. Pembrook
Title of Invention

Gas Analyzer,
Employing a Grating
and a Gas Selector
Cell

Gas Analyzer
    Patent Application
    Serial No.  and/or
Contractor Disclosure No.

        33-059
        (A-73-07)
        31-154-A
        (A-71-29)
Philco-Ford is currently requesting that the Government waive title and
retain only a non-exclusive royalty-free, irrevocable license in order
that the Contractor may file for these inventions and pursue additional
in-house work to  further exploit their application in the public interest.
                                   -39-

-------
  BIBLIOGRAPHIC DATA
  SHEET
                      1. Report No.
                       EPA-650/2-73-030
            3. Recipient's Acceaaion No.
   Instrument  to Monitor CH^,  CO,  and C02  in  Auto Exhaust.
            J. Report Date
               October  1973
 7. Aixhord)
    D. R. Burch  and .T. n
            •• Performing Organisation Kept.
              No.
 9. Performing Organization N>me and Address
   Philco-Ford  Corporation
   Aeronutronic Division
   Ford Road
   Newport Beach,  California  92663
            10. Project/Task/Voilc Unit No
            II. Contract/Grant No.

                68-02-0587
 12. Sponsoring Organization Name and Addreis
   EPA,  Chemistry  and Physics Laboratory,
   National Environmental Research Center
   Research Triangle Park, North  Carolina 27711
            13. Type of Report it Period
               Covered

               Final Report
                                                                      U.
 IS. Supplementary Notes
  16. Abstracts
      An infrared analyzer employing  gas cell correlation techniques and
      thennoelectrically-cooled photodetectors has  been designed  and con-
      structed to measure the concentrations of methane,  carbon monoxide, and
      carbon dioxide  in automotive  exhausts.  The dynamic range from the mini-
      mum detectable  concentration  to the maximum concentration that can be
      measured accurately for each  gas  is:  methane,  0.3  ppm to 10,000 ppm,
      carbon monoxide,  0.5 ppm to 10,000 ppm, and carbon  dioxide,  10 ppm to
      250,000 ppm.  The wide dynamic  range is made  possible by employing two
      different sample  cell  lengths  for each gas.  The concentrations of the
      three gases are measured simultaneously and independently.   Discrimi-
      nation against other gases in the  automotive  exhaust is very good.
 17. Key Words and Document Analysis.  )7o. Descriptors
      Air  pollution
      Measuring instruments
         Infrared spectrophotometers
      Exhaust gases
         Methane
         Carbon dioxide
         Carbon monoxide
 7b. Identifiers/Open-Ended Terms

      Air pollutant measuring   equipment
      Gas correlation spectrophotometer
 7e. COSATJ Field/Group   13B, 7B,  7C
  I. Availability Statement
FONM NTIS-19 IREV. S-721
19. Security Class (This
   Report)
     UNCL.
                                                          2v> security
                                                              '
                                                              UNCL
                                                                   ASSIFIED
                                                                                21. No. of Pages
                                                                                   46
                                            -40-
                                                                               U1COMM-DC UII2-P71

-------