Unit«d Stitw
Environmental Prottction
Agtncy
Offici of Mann*
and Estuarin* Protection
Washington DC 20460
March 1987
EPA 430.9-86-004
W«t«r
Quality Assurance/Quality Control
(QA/QC) for 301 (h) Monitoring
Programs: Guidance on  Field
and Laboratory Methods

-------
EPA Contract No. 68-01-6939
TC 3953-04

Final Report
QUALITY ASSURANCE/QUALITY CONTROL (QA/QC)
FOR 301(h) MONITORING PROGRAMS:
GUIDANCE ON FIELD AND LABORATORY METHODS
for

Marine Operations Division
Office of Marine and Estuarine Protection
U.S. Environmental Protection Agency
Washington,  DC  20460
 March,  1987
 by

 Tetra Tech, Inc.
 11820 Northup Way,  Suite 100
 Bellevue, Washington  98005

-------
                                 PREFACE
     This manual  has been prepared  by EPA's  Marine Operations  Division,
Office of Marine and Estuarine Protection in  response  to requests from EPA
Regional Offices  and coastal  municipalities  with  sewage  treatment  plants
discharging into estuarine and marine  coastal waters.   The  members  of the
301(h) Task Force of EPA,  which includes representatives  for  the EPA Regions
I, II,  III,  IV,  IX and  X, the Office of Research  and  Development,  and the
Office of Water, are to  be commended  for their  vif   role in  the development
of this guidance  by  the  technical  support  contractor,  Tetra  Tech,  Inc.
Under regulations implementing Section 301(h) of the Clean Water Act, munici-
palities are required to conduct  monitoring programs to  evaluate the impact
of their discharge  on marine biota, to demonstrate compliance  with applicable
water quality  standards,  and to measure toxic substances in the discharge.
The  collection and analysis of  high  quality data  require  that  specific,
established quality  assurance  and quality  control  (QA/QC)  protocols  be
adhered to in each of these major  monitoring  programs.

     QA/QC procedures are included in this document for environmental variables
that may  be measured in  effluent, receiving  water, sediment,  and organism
tissues sampled  during  301(h)  monitoring  programs.   Quality assurance and
quality control procedures are provided for sample  collection, field sample
handling, and  laboratory processing to  implement  specific monitoring program
requirements provided in the 301(h) modified  NPDES  permit.

     This manual is the result of several  years of  experience and effort in
designing,  implementing,  and  reporting results  of field  and laboratory
monitoring programs with the avowed purpose  of encouraging  both national and
regional consistency in estuarine  and  marine monitoring  programs.  In this
regard, special appreciation  is  due to the members  of the  301(h) Task  Force
who  contributed their valuable  time and expertise in the  development of this
manual.

-------
     The information provided  herein will  be  useful  to U.S. EPA monitoring
program  reviewers,  permit writers,  permittees,  and other organizations
involved  in  performing nearshore  monitoring  studies.  As  the monitoring
variables  included  in  t*vs document are  commonly used in  many marine and
estuarine monitoring programs, the guidance provided herein has broad applica-
bility beyond the 301(h) program.

-------
                                  CONTENTS


                                                                         Page
LIST OF FIGURES                                                            y

LIST OF TABLES                                                            vi

ACKNOWLEDGMENTS                                                          vii

INTRODUCTION                                                               ^

     PURPOSE                                                               j

     SCOPE                                                                 !

     FORMAT                                                                2

EFFLUENT MONITORING                                                        7

     GENERAL METHODS                                                       7

          Sampling Preparation                                             7
          Sampling Procedures                                              8
          Sample Handling                                                   g

               Field Procedures                                             g
               Sample Shipment                                            15
               Laboratory  Procedures                                      19

     EFFLUENT ANALYSES                                                    22

          Flow                                                            77
          PH                                                              g
          Temperature                                                     28
          Turbidity                                                       30
          Total  Suspended  Solids                                           34
          Settleable Solids                                                37
          Floating  Particulates                                            39
          Dissolved Oxygen (Winkler  Method)                                41
          Dissolved Oxygen (Probe Method)                                  44
          Biochemical  Oxygen  Demand  (BOD)                                  47
          Total  Chlorine Residual                                          50
          Oil  and Grease                                                   52
          Nitrogen  (Ammonia)                                               55
          Nitrogen  (Total Kjeldahl)                                        57
          Nitrogen  (Nitrate-Nitrite)                                       59
          Phosphorus  (Total)                                               61
          Priority  Pollutant Metals                                        63
          Priority  Pollutant Organic Compounds                            68
          Total  and  Fecal Coliform Bacteria                                74
          Enterococcus Bacteria                                            77


                                 11

-------
MONITORING THE RECEIVING ENVIRONMENT                                      80

     GENERAL METHODS                                                      80

          Sampling Preparation                                            80
          Sampling Procedures                                             83

               Station Location                                           83
               Water Sampling                                             84
               Grab Sampling                                              90
               Trawl Sampling                                             97

          Sample Handling                                                103

               Field Procedures                                          103
               Sample Shipment                                           113
               Laboratory Procedures                                     114

          Shipboard Laboratory Analyses                                  117

     RECEIVING WATER ANALYSES                                            119

          pH                                                             120
          Salinity                                                       124
          Temperature                                                    127
          Color                                                          129
          Transparency                                                   131
          Turbidity                                                      132
          Total Suspended Solids                                         135
          Settleable So .lids                                               138
          Floating Particulates                                          140
          Dissolved Oxygen (Winkler Method)                              142
          Dissolved Oxygen (Probe Method)                                145
          Biochemical  Oxygen Demand (BOD)                                148
          Oil  and Grease                                                 151
          Nitrogen (Ammonia)                                             155
          Nitrogen (Total  Kjeldahl)                                      157
          Nitrogen (Nitrate-Nitrite)                                     159
          Phosphorus (Total)                                             161
          Total and Fecal  Coliform Bacteria                              163
          Enterococcus Bacteria                                          166
          Chlorophyll  a                                                  169
          Phytoplankton                                                  172

     SEDIMENT/INFAUNA  ANALYSES                                            177

          Grain Size                                                     178
          Total Solids/Water Content                                     181
          Total Volatile Solids (TVS)                                     183
          Total Organic Carbon  (TOC)                                     185
          Biochemical  Oxygen Demand (BOD)                                187
          Chemical  Oxygen  Demand (COD)                                   190
          Oil and Grease                                                 192


                                 111

-------
          Sulfides (Total  and  Water  Soluble)                             196
          Priority Pollutant Metals                                      199
          Priority Pollutant Organic Compounds                           205
          Infauna                                                       212

     BIOACCUMULATION/TRAWL ANALYSES                                      234

          Priority Pollutant Metals                                      235
          Priority Pollutant Organic Compounds                           243
          Demersal Fishes  and  Megainvertebrates                         251

REFERENCES                                                              255

GLOSSARY                                                                260

APPENDIX A     RECOMMENDED METHODS FOR METALS IN EFFLUENT               A-l

APPENDIX B     RECOMMENDED METHODS FOR ORGANIC  COMPOUNDS IN EFFLUENT    B-l
                                  1V

-------
                                  FIGURES

Number                                                                  Page
   1    An example of chain-of-custody record                            16
   2    Examples of a sample analysis request form and a shipping  seal    17
   3    Deployment of a grab sampler                                     93
   4    Examples of acceptable and unacceptable grab samples              95
   5    Transect length as a function of  urrent speed                  101
   6    An example of a chain-of-custody record                         111
   7    Examples of a sample analysis request form and a shipping  seal   112
   8    Construction of a sieve box                                     213
   9    Example of a sieving stand                                      218

-------
                                  TABLES
Number                                                                  Page
   1    Environmental variables  included  in  this  document                  3
   2    Recommended  sample  sizes, containers,  preservation,  and
        holding  times for effluent  samples           '                     10
   3    Recommended  methods  for  measuring effluent  variables             12
   4    Examples of  problems  frequently  encountered during offshore
        surveys  and  possible  solutions to each problem                    82
   5    Recommended  sample  si':es, containers,  preservation,  and
        holding  times for offshore  samples                               104
   6    Recommended  methods  for  measuring offshore  variables            106
   7    Some  common  external  abnormalities observed in  fishes  from
        polluted areas                                              .     254
  A-l   List  of  approved  inorganic  test  procedures  [note:  this  is a
        reproduction of Table IB of U.S. EPA (1984)]                    A-l
  B-l   List  of  approved  test procedures  for non-pesticide organic
        compounds [note:  this is a reproduction  of Table 1C of
        U.S.  EPA (1984)]                                                 B-l
  8-2   List  of  approved  test procedures  for pesticides  [note:
        this  is  a reproduction of Table  ID of  U.S.  EPA  (1984)]           B-3
                                vl

-------
                             ACKNOWLEDGMENTS

     This document has  been  reviewed by the 301(h) Task Force of the U.S. EPA,
which Includes  representatives  from the Water Management Divisions of U.S. EPA
Regions I, II, III,  IV,  IX, and X; the Office of Research  and Development
- Environmental Research Laboratory - Narragansett (located  in  Narragansett,
RI and  Newport,  OR),  and the Marine  Operations Division  in  the Office of
Marine, and Estuarine Protection,  Office of Water.   Among  the reviewers,
the assistance provided by Dr. Steve Ferraro under the direction of Dr. Donald
J. Baumgartner, Dr.  Dona".  Phelps  under the direction of Dr. Allan Beck,
and Dr. Brian Melzian  under  the direction of Patricia Eklund  in coordinating
the numerous technical  comments from the Office of Research  and Development
and Region IX is gratefully  acknowledged.

     This  technical guidance  document was prepared by Tetra  Tech Inc. for
the U.S. EPA Operations Division, Office of Marine and Estuarine Protection,
Office  of Water) under the  301(h) post-decision technical  support contract
No. 68-01-6938, Allison J.  Duryee,  Project  Officer.  This  document was
prepared under the direction of Dr.  Thomas Ginn  (Program  Director) and
Dr. Scott Becker (Work  Assignment Manager) of Tetra Tech, Inc.  The following
Tetra Tech staff members contributed to specific sections:  Ms. Ann Bailey
(effluent variables,  priority pollutant metals), Mr. Robert Barrick (priority
pollutant organic  compounds), Dr. Scott Becker (effluent  variables, water
column variables, demersal fishes and megainvertebrates, conventional sediment
variables), Dr.  Gordon Bilyard (infauna, conventional sediment variables),
and Ms.  Julia Wilcox (priority pollutant organic compounds, priority pollutant
metals).  The following  outside consultants also contributed to specific
sections:  Mr. Jack Word (Evans Hamilton, Inc. - infauna), Dr. Victor  Cabelli
(CABs Associates - microbiological  variables),  Ms. Kathy  Krogslund and
Mr. Ron Citterman (University of Washington - water column variables).
                                   vll

-------
                               INTRODUCTION
PURPOSE

     This document was produced  in response to specific regional  requests
for assistance on  technical issues raised during issuance of 301(h)-modified
NPDES  permits.  This  technical  g-idance to regional program offices and
permittees provides the framework for making informed decisions with respect
to  the field and  laboratory methods  used when  moo toring the effects of
sewage  discharge on marine and estuarine ecosystems.  Tht principal objectives
of this document are to ensure that:

     •    Samples are collected, processed, stored, shipped, and analyzed
          using acceptable and standardized procedures

     •    Quality of generated data  is documented adequately

     •    Results are reported completely  and correctly

     •   Security of samples and data  is  maintained  at all times.

SCOPE

     The information  presented in  this  document  is designed to complement
the sampling and analysis specifications contained  in the permits  for  individual
monitoring  programs.  Those specifications  identify the  sampling  strategy
of each survey {e.g., station  locations,  field  replication,  time of  sampling),
as  well  as  which environmental  variables  will  be  measured.   It  is  likely
that different monitoring programs will  require different combinations
of variables to be measured, depending  upon site-specific  and  discharge-
specific considerations.  In this  document,  collection  and  analysis  procedures

-------
are  presented  for  most environmental  variables  that may  be  measured  in
effluent, receiving water,  sediment,  and tissue during a 301(h)  monitoring
program (Table I).

      Information presented  in this document can be  used to address most
of the quality assurance/qual ity  control  (QA/QC) plan elements described
in  "Guidance  for Preparation of  Combined  Work/Quality Assurance Project
Plans for Environmental Monitoring"  (U.S. EPA  1984).  In the present document,
guidance is given for the  following activities:

      t    Preparation for  sampling
      •    Sample  collection
      •    Sample  processing
      •    Sample  size
      •    Sample  containers
      •    Sample  preservation
      •    Sample  holding times
      •    Sample  shipping
      •    Logkeeping
      •    Labeling
      •    Custody procedures
      •    Analytical methods
      •    Calibration and preventive maintenance
      •    Quality control checks
      •    Corrective action
      •    Data reporting requirements.

FORMAT

      This document is  divided into  two major sections:  Effluent Monitoring
and Monitoring of the Receiving Environment.  This division was  used  because
 these two kinds of monitoring  typically are  not  conducted by the same organ-
 ization.  For example, effluent sampling frequently is conducted  by treatment
 plant  personnel,  whereas sampling/of the receiving environment generally
 Is conducted by outside contractors.  Because of this possible  separation,
each major section of this report  is designed to stand alone.  Thus,  effluent

-------
TABLE 1.  ENVIRONMENTAL VARIABLES INCLUDED IN THIS DOCUMENT
Matrix
Variable
Flow
PH
SaMnity
Temperature
Color
Turbidity
Transanssivity
Transparency
T' :al suspended solids
Se .tleable solids
Floating participates
Dissolved oxygen
Biochemical oxygen demand
Chemical oxygen demand
Total chlorine residual
Oil and grease
Nitrogen (ammonia)
Nitrogen (total Kjeldahl)
Nitrogen (nitrate-nitrite)
Phosphorus (total)
Sul fides
Priority pollutant metals
Priority pollutant organic compounds
Total and fecal col i form bacteria
Enterococcus bacteria
Chlorophyll £
Phy to plankton"
Grain size
Total solids
Total volatile solids
Total organic carbon
Infauna
Demersal fishes and
epibenthic macrolnvertebrates
Effluent
X
X

X

X


X
X
X
X
X

X
X
X
X
X
X

X
X
X
X








Receding
Water

X
X
X
X
X
X
X
X
X
X
X
X


X
X
X
X
X



X
X
X
X







Seabed/
Sediment Tissue












X
X

X

»


X
X X
X X



X
X
X
X '
X

X

-------
sa.-nplers will  not  have to si^t through information relevant only to those
who sample the receiving environment and vice versa.  Although  this  independence
of  sections  leads  to  some redundancy,  its overall  efficiency outweighs
this drawback.

     Within the effluent  and  receiving-environment monitoring sections,
collection  and analysis  procedures are presented first for  general sampling
methods and  then  for  specific  environmental  variables.   General sampling
method  sections include procedures  *or 1) preparing for a survey, 2) collecting
t^e  samples  from  which subsamples  will be taken  for  specific variables
(e.g.,  effluent sampling, water  sampling, grab  sampling, trawl  sampling),
and  3) handling  samples  in  the  field (labeling, chain-of-custi ly, storage,
shipping) and  in the laboratory  (reception,  tracking, log  keeping).   Each
section on  sample handling  also  contains  tables that summarize the recommended
collection  specifications  and analytical methods for each environmental
variable.

     The collection and  analysis procedures for  specific environmental
variables are organized to comply with most  of  the "elements of  a QA project
plan"   described  by U.S.   EPA (1984).  The  following elements  are addressed
for  each variable:

      1.  Field collection

     2.  Field processing

      3.  Analytical procedures

      4.  Calibration and  preventive maintenance

      5.   Quality  control  checks

      6.   Corrective action

      7.   Data quality  and reporting.

-------
     1.)  The section on field  collection  describes the recommended  container
type, special cleaning procedures, and special  collection  techniques for
each variable.

     2.)  In the field processjng section,  recommendations  are  made for
cne preservation technique  and recommended  maximum holding  time before
analysis.

     3.)  The section on  analytical procedures identifies  the method(s)
recormended for laboratory analysis for each  variable.  The methods recommended
in this document take precedence over those  identified in Tetra Tech (1985c).
To conform with U.c. EPA's equivalency policy, multiple methods were recommended
for effluent and receiving water variables if  they have been shown to produce
equivalent  results  (U.S.  EPA 1984).  Multiple methods  were confined  to
tnose  approved by  U.S. EPA  (1979b, 1984)  and the American Public Health
Association  (APHA 1985).   Although other methods may  provide equivalent
results (e.g., American  Society for Testing  and Materials, U.S. Geological
Survey) they generally were not  recommended.  However,  if no U.S. EPA-
or APHA-approved methods were available for  a particular  effluent  or receiving-
'*ater variable, an alternate source of methods was used.   Because  no U.S. EPA-
or  APHA-approved methods are currently available  for sediment and tissue
variables, alternate  sources of methods were  used for  all of these variables.
 In  all cases,  priority  was given to methods that were developed under the
guidance of U.S. EPA (e.g., Plumb  1981; Tetra Tech 1986a,b).   Methods  for
 collection and analysis  of benthic infauna were described  in  considerably
 greater detail  than  those for other variables because of the importance
 of these organisms as  indicators of sewage-related biological impacts and
 because of  the  lack of a published reference having adequate depth and
 scope.   In addition to  specifying  the  recommended  analytical method(s),
 the  section  on analytical procedures describes the  major sources of interference
 or error  that  may  be encountered when using the  methods.   This kind  of
 information was included  because  it often is  incorporated  into the method
 description without being highlighted and because some methods do  not  treat
 it as comprehensively as  others.

-------
     4.)   The section  on  cal Ibration and  preventive maintenance describes
the kind and frequency  of calibration and  equipment maintenance that are
essential to providing  quality data for  each variable.

     5.)   In the  section  on quality control  checks, the kind and frequency
of quality control checks (e.g., duplicates,  blanks, spiked samples, reference
material analyses)  that  should be conducted  during  laboratory analyses
are described for  each  variable.  A duplication level  of  10 percent was
considered appropriate for most  variaoles.   Blanks and  spiked  samples
generally were recarmended at a frequency  of  one per batch.  It is recommended
that duplicate analyses  be distributed  relatively evenly throughout the
full sequence of samples analyzed, rather  tr in  being concentrated at the
beginning or end of the sequence.

     6.)   The section on corrective action describes major sources of analytical
errors that may be encountered when analyzing for each  variable. Possible
solutions also are recommended.

     7.)   The section  on  data qua! ity  and reporting  describes documented
levels  of accuracy  and precision (if available) achieved  using the analytical
method,  the kinds  of  information that should  be reported to U.S. EPA, and
the units and significant figures that should be  used when reporting results.
Specifications in  this section are compatible with the data requirements
of the 301(h) program.   Further guidance on the  data reporting requirements
is  available in the "GOES Data Submissions Manual"  (Tetra Tech and American
Management Systems 1985).

     Because this document  considers a wide range  of  disciplines, the term-
inology  used  in certain sections may  be  unfamiliar  to various  readers.
Therefore, a glossary is provided  at  the end of the document to define
specialized words or phrases.

-------
                           EFFLUENT MONITORING
     Recommended  methods for measuring  effluent  variables during  301(h)
monitoring programs are described  in  this section.  The initial major  section
includes general  procedures  for pre-sampling  activities,  sampling, and
sample handling.   The  second major section  presents detailed procedures
for measuring 20 effluent variables.  The  general  section on sample handling
contains tables that summarize the  recommended collection  specifications
and  analytical methods  for the spe  ific effluent variables.   Tables  of
recommended methods  for metals and organic compounds  in  effluent are also
presented in the appendix.

GENERAL  METHODS

Sampling Preparation

      The QA/QC Project  Plan  should be  thoroughly reviewed  in  advance by
 sampling personnel to  identify the  following:

      t    Responsibilities for  each individual  associated with the
          sampling and analysis

      •    Statement and prioritization of study objectives

      •     Background  information, sampling locations, sampling frequency,
           and  sampling procedures

      •     Variables  to be  measured and corresponding required  sample
           sizes,  containers preservatives, and holding times

      •     Sample splits or performance samples to be submitted with
           the  samples

-------
     •    Laboratories to -which samples will  be shipped

     •    Shipping requirements.

     The  study objectives and  their prioritization should be understood
by all personnel involved in the  sampling effort.   This  will ensure  that
if modifications  of the plan become necessary, their  impact on the overall
goals of the sampling effort can be adequately evaluated.

     i'o ensure that all  required  sampling equipment and  supplies are available
at the time of sampling  an equipment checklist  should be constructed.
Spare  parts and  backup supplies  (e.g., extra glass  jars,  spare pH probe)
should be included in the inventory.

Sampling Procedures

     Sample collection equipment, sampling frequency,  and sample locations
depend upon the size and nature of  the discharge and generally are specified
in  the discharge permit.   Decisions regarding equipment to be used during
the monitoring program  should be  made after  reviewing related  literature
(e.g., Tetra  Tech 1982;  U.S. EPA 1982).   If  automatic samplers are used
for collection of flow-proportioned 24-h composite samples, the  sampler
should be selected on the basis of  the analyses to be performed.  The automatic
sampler must be able to collect adequate  sample sizes without  introducing
contaminants.  For example, plastic tubing  and bottles cannot  be  used  to
collect  samples for priority pollutant  organic  compounds.

      Grab samples, rather  than  composite samples, must be taken  for  certain
variables (e.g., dissolved gases, volatile  compounds, microbiological  variables)
because  of  changes that are likely  to occur during storage.  Effluent variables
that  must be collected as  grab  samples  are pH,  temperature, total  and fecal
coliform bacteria, dissolved  oxygen,  oil  and grease, and volatile  organic
compounds.   For further sample collection  requirements,  the  section  on
 field collection should  be consulted  for each variable.  Note that  certain
variables (e.g., volatile organic compounds) require  filling sample containers
 completely,  leaving no air  space, whereas other variables  (e.g.,  total
                                      8

-------
and fecal  coliform,  oil  and grease) require leaving air space  (i.e., headspace)
to allow for adequate mixing or addition of acid.

Sample Handling

     Proper  sample handling  ensures that  changes  in  the  constituents of
interest are minimized  and guards against errors when collecting, shipping,
and analyzing samples.  Recommended  sample size, container, preservation
and storage requirements  for each effluent variable  are summarized in Table 2.
Recommended laboratory methods  for  each effluent variable are listed in
Table 3.  These  requirements are also noted in subsequent  sections.   These
requirements and  methods should be reviewed in advance j, laboratory personnel
to ensure that the sample size, containers,  preservatives,  and all  other
specifications are consistent  with the needs and  capabilities of the laboratory.

Field Procedures-

     It is important throughout any sampling  and analysis program to maintain
the integrity of the sample from  the time of collection  to the  point of
data  reporting.   Proper chain-of-custody  procedures allow the possession
and handling of samples to be  traced  from collection  to  final disposition.
It  is  recommended that chain-of-custody procedures be used for all sampling
conducted during 301(h)  monitoring programs.  The  documents needed to maintain
proper  chain-of-custody include:

     t     Field logbook — All pertinent  information  on  field activities
           and sampling efforts should  be  recorded  in a bound  logbook.
           The field supervisor is  responsible for ensuring that  sufficient
           detail  is recorded in the .logbook.   The  logbook  should enable
           someone else  to  completely reconstruct the  field  activitiy
           without relying on the memory  of the field  crew.   All  entries
           should be made in  indelible  ink, with each  page signed  and
           dated by the  author, and a  line drawn through the remainder
           of any page.  All  corrections  should  consist of permanent
           line-out deletions that are  initialed by  the field  supervisor.
           At a minimum, entries in a  logbook should include:

-------
    TABLE 2.   RECOMMENDED  SAMPLE  SIZES, CONTAINERS, PRESERVATION,
                 AND  HOLDING TIMES  FOR  EFFLUENT SAMPLES*
Minimum
Sample
Slzeb
Measurement («l)
pH 25
Temperature 1 .000
Turbidity 100
Total suspended solids 1,000
Settleable solids 1.000
f\? ing pa.-ticulates 5.000
Dissolved oxygen
Probe 300
Kinkier 300
Biochemical oxygen
demand 1 .000
Total chlorine
residual 200
Oil and grease I. 000
Nitrogen
Ammonia-N 400
Total Kjeldahl-N 500
Nitrate * Nitrite-N 100
Phosphorus (total) 50
Priority pollutant metals
Mercury 100
Metals, except mercury 100
Priority pollutant
organic compounds
Ex tractable compounds 4,000
(includes phthalates,
nitrosamines, organo
rhlnrlnp ne«Meide«.
ContalnerC
P.6
P.6
P.S.
P.G
P.G
P.G
G bottle & top
6 bottle ft top
P.G
P.G
6 only
P,G
P.G
P.G
P. 6
P.G
P.G
6 only.
TFE- lined
cap
Preservative^
None
None
Cool. 40c
Cool, 40c
Cool. 4oc
None
None
Fix on site; store
in dark
Cool. 40c
None
Cool, 40c
H2S04 to pH<2
Cool , 40c HjSOa to
pH<2
Cool . 4»c H2S04 to
pH<2
Cool . 4% H2S04 to
PH<2
Cool. 40c H2SO« to
pH<2
HN03 to pH<2
HN03 tc pH<2
Cool . 40 c
0.0081 NajSzOjg
Store in dark
Maiimum
Holding
Time
Analyze immediately^
Measure immediately^
48 h
7 days
48 h
Analyze Immediately6*
Analyze immediately6
8 h
48 h
• Analyze immediately0
28 days
28 days
28 days
28 days
28 days
28 days
6 mo
7 days until
extraction
40 days after
extraction
  PCBs, nitroaromatics,
  isophorone, polynuclear
  aromatic hydrocarbons,
  haloether, chlorinated
  hydrocarbons, phenols,
  and TCOD)
Purgeable compounds
40
G only, TfE-  Cool, 4»C
lined  septum  0.0081
7 days"
                                        10

-------
TABLE  2.   (Continued)
Total and fecal
conform bacteria 250-500
Enterococcus bacteria 250-500

P. 6 Cool. 40 c
0.0081 NajS^Q
P. 6 Cool, 40 C
0.0081 Ma2S203g
6 h

6 h

  a Reference:   Adapted from U.S.  EPA  (1979b.  1984).

  0 Recommended  field sample sizes for one laboratory analysis.   If additional laboratory analyses
  are required  (e.g.. replicates), the field sample size should  be  adjusted accordingly.

  c Polyethylene (P) or Glass (6)

  
-------
TABLE 3.  RECOMMENDED METHODS FOR MEASURING EFFLUENT VARIABLES
Method Reference
Variable
pH
Temperature
Turbidity
Total suspended solids
Settleable solids
Floating particulates
Dissolved oxygen
Probe
Uinkler
Biochemical oxygen demand
Total chlorine residual
Oil and grease
Nitrogen
Ammonia-N
Total Kjeldahl-N
Nitrate-t-Nitrite-N
U.S. EPAa
150.1
170.1
180.1
IjQ.Z
160.5
-
360.1
360.2
405.1
330.1
330.2
330.3
330.4
330.5
413.1
350.1
350.2
350.3
351.1
351.2
351.3
351.4
353.1
353.2
353.3
APHAb
423
212
21 4A
214B
209C
209E
206Ad
421F
421B
507
408A
408B
408C
4080
408E
503A
41 7A
417B
4170
417G
417B
417D
417E
420A
4208
418C
41 8F
Otherc
-
-
-
-
-
-
-
-


-

-
                                 12

-------
TABLE 3.  (Continued)
Phosphorus (total)
Priority pollutant metals
Priority pollutant
Organic compounds
Total coliform bacte -ia
Fecal coliform bacteria
Enterococcus bacteria
365.1
365.2
365.3
365.4
Table 1B&
U.S. EPA
(1934)
Table ice
U.S. EPA
(1984)
-
-
-
42 4C
42 4F
42 4G
Table IBS
U.S. EPA
(1984)
Table ice
U.S. EPA
(1984)
908Af
909A9
908Cf
909C9
-

™

p. 114f,h
p. lOSg.h
p. 132f,h
p. 1249,h
U.S. EPA1
a Methods recommended in U.S.  EPA (1979b).
D Methods recommended in APHA (1985).
c Methods recommended in sources other than U.S.  EPA (1979)  or APHA (1985)
when  no methods were recommended in the latter two  sources.
d This method  is tentatively recommended by APHA.
e The list  of U.S. EPA  and  APHA methods for individual  components of this
group are listed in the table specified and  are too extensive to  include
here.
f This method  can be used whether or not chlorine is present.
9 This method  can be used only when chlorine  is absent.
h Page reference of this method  in Bordner et al. (1978).
 i U.S. EPA  is  currently finalizing a recommended method for this variable.
                                       13

-------
    Date and time of starting work

    Names of field supervisor and team members

    Purpose of proposed  sampling effort

    Description of sampling  site,  including information
    on any photographs that  may be  taken

    Location of sampling site

    Details of actual  sampling effort, particularly deviations
    from standard operating  procedures

    Field observations

    Field measurements made  (e.g.,  pH, temperature, flow)

    Field laboratory analytical  results

    Sample  Identification

    Type and number of samples collected

    Sample  handling,  packaging, labeling, and  shipping
     information  (including destination).

Chain-of-custody procedures  should be  maintained with the
field  logbook.  While  being  used  in the  field,  the logbook
should remain with the field team at all times.  Upon  completion
of the  sampling effort,  the logbook should be kept  in a
secure  area.   All  logged  information  should be summarized
and submitted  to  U.S.  EPA after sampling  is completed.
                            14

-------
     t     Sample  labels -- Sample labels  must  be  waterproof and must
          be securely fastened to the outside of  each  sample container
          to  prevent misidentlfication of samples.  Labels  should
          contain at least  the sample number, preservation technique,
          date  and  time of collection,  location  of collection, and
          signature of the  collector.  Labels should be marked  with
          Indelible Ink  and placed on the  body of the  jar.  Abbreviated
          labels may also be placed on the  cap of  each  jar to facilitate
          sample Identification.

     t     Chain-of-custody  records -- A chain-of-custody record (Figure 1)
          must  accompany every group   f samples.   Each  person  who
          has  custody  of  the  sample onst  sign the  form and  ensure
          that  the  samples are not  left unattended unless  secured
          properly.

     •     Custody seals -- Custody seals (Figure 2) are used to detect
          unauthorized tampering with the samples.  Sampling personnel
          should  attach seals to  all  shipping  containers sent to the
          laboratory by common carrier.  Gummed paper seals or  custody
          tape  should.be used so  that the  seal must be broken when
          the container is  opened.

For further information regarding proper chain-of-custody procedures, consult
the policies and  procedures manual for the National Enforcement Investigations
Center  (NEIC; U.S. EPA 1978).

Sample  Shipment-

     All  preserved  samples  should  be shipped  immediately after  completion
of sampling.  This minimizes the  number of people  handling  samples, and
protects sample quality and  security.   Guidance for shipping  hazardous
materials can be  found  in  U.S.  Department  of Transportation  (1984).  As
samples are prepared  for shipping,  the  following  should be kept  in mind:
                                     15

-------
Uniud SutH
Environment* Protection
Agency
RtponlO
1200 SlJrth AOTMM
$••«)• WA 98101
CHAIN Of CUSTODY RECORD
PROJECT
LAB *











STATION











DATE











TIME











RELINQUISHED BY: O««M»
RCUNOUI8HED BY: «*•«•»
REUNQUI8HED BY: /S*«M
RELINQUISHED BY: '!»—*
DISPATCHED BY: «w— *
METHOD OF SHIPMENT:

DATA
SAMPLERS: <&?•*•
SAMPLE TYPE
I
m
m











a











1











5











9











s
a











n
5











REMARKS











MCIIVtD BY: ft+~*
RECEIVED BY: av^um
RECEIVED BY: ts*~~
RECVO BY MOBILE LAB FOR WELD
ANAL.: is****
fTIMI
RECEIVED FOR LAB BY: r*v«—
DATEH1MI
DATE/TIME
DATE/TIME
DATimMB
DATE/TIME

Ontnbuoon
        On* Coo»-Surv«v Coaidniiar ftta f»m
        Figure 1.  An example of  * chain-of-custody record.  An equiva-
                    lent form should be used  for 301(h) monitoring.
                                         16

-------
    ENVIRONMENTAL PROTECTION AGENCY
CLP Sample Management OH ice
P.O. Boa SIS • Alexandria. Virginia 22313
Phone:  703/J37-2MO - FTS/3J7-2»90
SAS Number
                       SPECIAL ANALYTICAL SERVICE
                              PACKING LIST
Sampling Officei
Sampling Contacts
(name)
(phone)
Sample
Number*
Sampling Datefsh
Date Shipped:
Site Name/Code!

Ship Tm
AttK
For Lab UM Only
Date Sample* Rec*di
Received By:

Sample Description Sample Condition on
LA, Anaiyiis, UatrU, Concentration Receipt at Ub
 1. _
 2. .
 J. .
 a. _
 J. .
 4. .
 7. .
 S. .
 ». ,
10. .
II..
12- .
13. .
!*•.
I'-
ll
17.
'«•
19.
20.
                                                            For Ublte Only
                           EPA CUSTODY SEAL
  KNYiROMUMTAL PtOTICTMN ACtNCY
                                Meri MM Afl» 71 TUB «•<
             *
            si
    Figure  2.  Examples  of a  sample  analysis  request  form (above)
                and a custody  seal  (below).  Equivalent material
                should be used for  301(h)  monitoring.
                                    17

-------
•    Shipping containers  should  be in good  shape  and capable
     of withstanding rough treatment  during  shipping.

•    Samples  should be packed tightly:

          Dividers must separate all  glass containers

          Empty space within  shipping  boxes  should  be filled  so  that
          jars are held securely.

•    All containers  must be leak-proof.   If a container is not
     leak-proof by design, the interior   should be  lined with
     two  heavy-duty plastic  bags  and  the tops of  bags  should
     be tied  once  samples are inside.  Adequate absorbent material
     should  be placed  in the  container in a  quantity sufficient
     to absorb all of the liquid shipped.
      •
t    All samples  should be accompanied by a  sample analysis request.
     Variables to  be  analyzed by the laboratory,  and total  number
     and  kind of  samples shipped for  analysis should be listed
     on the request sheet.   An  example  sample  analysis  request
     form is  illustrated  in  Figure  2.  The laboratory  should
     acknowledge receipt  of shipment  by signing and  dating the
     form, and returning  a copy to the designated QA coordinator.

 •    A  chain-of-custody  record for  each shipping container should
     be filled out completely  and  signed.

 •    The  original chain-of-custody record  and analysis  request
     should be protected  from damage  and  placed inside  the  shipping
     box.   A copy  of each should  be  retained by  the shipping
     party.

 •    The  custody seal  should  be attached so  that  the  shipping
     box  cannot be opened without  breaking  the seal.
                                 18

-------
     •    For  shipping containers  carrying  liquid samples:

              A  "This  End  Up" label  should be attached to each  side
              to ensure that  jars are  transported in upright position

              A  "Fragile-Glass"  label  should be attached to the top
              of the box to minimize agitation of samples.

     •    Shipping containers  should  be  sent  by a carrier that  will
          provide a delivery  receiot.  This will  confirm  that  the
          contract  laboratory received the samples  and serve  as a
          backup  to the chain-of-custody  record.

     •    All  shipping charges  should  be prepaid  by the  sender  to
          avoid confusion and possible  rejection  of  package by the
          contract laboratory.

Laboratory Procedures--

     At the laboratory, one  person should be designated custodian of all
incoming samples.  An alternate should  also be  designated  to serve In  the
custodian's absence.  The custodian should  oversee the  following activities:

     •    Reception of samples

     •    Maintenance of chain-of-custody records

     •    Maintenance of sample tracking  logs

     •    Distribution of samples for laboratory analyses

     t    Sending samples to outside laboratories
                                     19

-------
     t    Supervision  of labeling,  log keeping,  data  reduction, and
          data transcription

     •    Storage and  security of all samples, data, and  documents.

     Upon  reception of  samples, a designated laboratory person should fill
out the chain-of-custody record,  indicating time  and  date of reception,
number  of samples,  and  condition of samples.  All  irregularities indicating
that sample security or  quality may hav«» been jeopardized (e.g.,  evidence
of tampering, loose  lids, cracked jars) should be noted  on the sample analysis
request form  and  returned to the  client-designated QA :oorrfinator.   In
addition, a designated  person  should initiate and maintain a  sample-tracking
log that will follow each sample through all stages of laboratory processing
and analysis.  Minimum information required  in a sample-tracking log includes:

     •    Sample identification number

     t    Location and condition of storage

     •    Date and time of each removal  from and return to  storage

     t    Signature of person  removing and  returning the sample

     •    Reasons for  removal  from  storage

     •     Final disposition of sample.

     All  logbooks, labels, data  sheets,  tracking  logs, and custody  records
 should  have  proper  identification numbers and be filled  out accurately.
 All  information should  be written in ink.   Corrections should  be  made by
 drawing a  line through  the  error and  entering the correct  information.
 Corrections  should be  signed  and dated.   Accuracy of all data  reductions
 and  transcriptions should be verified at least  twice.   All samples and
 documents  should be  properly stored  within the laboratory until  the client
 authorizes  their removal.  Security  and  confidentiality of all stored material
 should be  maintained  at  all  times.   Before  releasing analytical results,

                                      20

-------
a',1  information on sample  tags, data  sheets, tracking  logs, and custody
records should be cross-checked to  ensure that  data pertaining  to each
sample are consistent throughout the record.

     Originals of  the following documents should be sent  to  the client:

     t    Chain-of-custody records

     t    Sample-tracking logs

     •    Data report sheets

     •    Quality  contiol records.

Copies of all forms should be retained by the laboratory in case originals
are  lost  in transit.

     When replicate  analyses are  conducted  as QA/QC checks, it is  recoimended
that they be distributed relatively  evenly throughout the  full  sequence
of samples analyzed,  rather than being  concentrated at some point  (e.g.,
beginning, end) of the sequence.  This  precaution will enhance the  probability
that if problems develop within part  of a sequence, they will be detected.
                                      21

-------
EFFLUENT ANALYSES

      In  this section, QA/QC  procedures  are presented for  the  following
20 effluent  variables:

      •     Flow
      •     pH
      •     Temperature
      •     Turbidity
      •     Total suspended solids
      •     Settleable solids
      •     Floating particulates
      •     Dissolved oxyg:r. (Winkler metiiod)
      t     Dissolved oxygen (probe method)
      •     Biochemical  oxygen demand (BOD)
      •     Total chlorine residual
      •     Oil and grease
      •     Nitrogen (ammonia)
      •     Nitrogen (total Kjeldahl)
      •     Nitrogen (nitrate-nitrite)
      •     Phosphorus (total)
      •     Priority pollutant metals
      •     Priority pollutant organic compounds
      •     Total and fecal coliform bacteria
      •     Enterococcus bacteria.
                                       22

-------
                                                                 Effluent
                                                                     Flow
Flow
     Accurate  flow measurement Is essential for proper operation and control
of wastewater  treatment plants.  Flow determinations are used  for  a  variety
of purposes,  including 1)  provision  of data with  which plant operation
and performance can be evaluated, 2) provision of data  for long-term planning
of treatment  plant capacity relative to utilized  control  capacity, and
3) determination of compliance with permitted effluent mass-loading limits.

     Flow can be measured jri  situ using  \  variety  of methods.  The two
major categories of methods are direct-discharge  and velocity-discharge
(Metcalf and Eddy 1979).  The direct-discharge methods are used most frequently
for measuring  wastewater flows.  These methods relate  the rate of  discharge
to one  or  two easily measured variables and employ  such devices as weirs,
Parshall flumes, Venturi meters,  and magnetic  flow meters.  Accuracy of
these direct-discharge methods generally ranges  from +3 to +5  percent of
the flow rate  (Hinrichs 1979).  Any system that cannot measure wastewater
flow within +10 percent is considered unacceptable for NPOES  compliance
(U.S. EPA 1977).

     The preferred  location  for  a flow-measuring device is at  the end of
all treatment  processes and downstream  front all sidestreams.   If the  device
is located  at the head of the treatment process, the  flow must be corrected
for all  return streams that originally discharged upstream from  the flow-
measuring  device.   For continuous  flow measurements, a system including
the primary flow device, a flow sensor,  transmitting equipment, a  recorder,
and possibly a totalizer are  recommended.

     Flow-measuring  equipment  should  be routinely  inspected, calibrated,
and maintained to ensure that accurate  data  are produced.   At a minimum,
equipment  should be serviced according to  the  frequency  recommended  by
the manufacturer.  Servicing  should be  more frequent if  the equipment does
not appear to  be functioning  properly.   A  common cause of improper equipment
operation is accumulation of  foreign material  in or on equipment components.
                                     23

-------
                                                                   Effluent
                                                                      Flow
     Flow  should be reported  as millions of gallons per  day (MGD).  Where
continuous monitoring of an  effluent  is required, flew measurements should

include the total daily flow and  the  peak daily flow on a 24-h basis.

-------
                                                                   Effluent
                                                                         PH
£H

Field Procedures--

     Col 1ection--Samp1es  for pH determination should be collected in poly-
ethylene or glass bottles  having airtight screw caps.  Because pH is unstable
and cannot be preserved,  these samples should be analyzed as soon as possible
after collection.  Because  pH of waters not  -,t equilibrium with the atmosphere
may  change upon  exposure to the  atmosphere,  sample containers should be
completely filled and tightly sealed after collection.

     Prior to filling, each sample  bottle and cap should be rinsed thoroughly
with sample water.  This can be  achieved by  filling the  bottle  halfway,
sealing and shaking It, and rinsing the stopper as the wash water  is discarded.

     Process ing--Because pH cannot be preserved, samples should be analyzed
immediately after collection.  If a short delay occurs, the samples  should
be  stored in the  dark at 4° C and the storage time should be noted on the
log sheet.

Laboratory Procedures--

     Analytical Procedures—Analytical  procedures are given  in U.S. EPA
Method 150.1 and APHA Method 423.   As noted  previously,  pH samples  should
be  analyzed as soon as possible  following  collection.

      Several  potential  sources  of interference with pH measurements should
be  avoided.  Because the response of the  electrode  can  be impaired if  it
 is  coated with oily or particulate material,  the  electrode should  be gently
blotted or washed with a detergent periodically.   Treatment  with  hydrochloric
acid  may  be necessary to remove  some  kinds of film.  Temperature can influence
pH  measurements by altering electrode output and  by changing the pH inherent
 in  the  sample.  The  first source of temperature interference can be controlled
by  using  a pH meter  having temperature  compensation  or by calibrating the
                                     25

-------
                                                                  Effluent
                                                                       pH
meter  at the temperature of  the samples.   Because the second  source of
temperature interference  cannot be controlled, the temperature  at which
the  pH determination of each,  sample  is  made  should be logged  and  reported.
It  is  recommendec that  sample temperature  never differ by more  than 2° C
from that of the buffer solution.

     When  pH  measurements are being  made,  it  is critical  that the sample
be stirr-Kl at a constant rate to provide  drift-free (<0.1 pH units)  measure-
ments.  The rate of stirring  should min-mize air  transfer at the surface
of the sample.   At  least 30 sec should be allowed for  each measuremen
to stabilize.

     Calibration and Preventive Maintenance—Calibration procedures should
follow specifications given by the manufacturer  of the  pH meter.    General
guidelines are given  by U.S. EPA (1979b)  and APHA (1985).

     Primary buffer  salts should be those of the National  Bureau of Standards
(NBS).   Secondary buffer salts can be prepared from the NBS  salts or purchased
as  a solution  that  has  been calibrated  to NBS salts.   The pH meter should
be  calibrated at a minimum of two points  that bracket the  expected  pH value
of  the  samples and  that are three or more pH units  apart (U.S. EPA 1979b).
Prepare  fresh  buffer solutions  at  least every month  to avoid erroneous
calibration due  to mold growth  or contamination.

     Preventive maintenance  procedures should follow  specifications given
by  the manufacturer  of the pH meter.   In general, verification of  electrode
performance and meter performance is  the only operator  service recommended.
An  electrode  should  be replaced when it  no  longer meets span  requirements
and does not  improve with rejuvenating procedures.

     Quality Control Checks—The  pH meter should be calibrated  at the beginning
of  each  series of samples  and after each  group of 10 successive measurements.
 It  is recommended  that duplicate  pH determinations  be made on at least
                                      26

-------
                                                                  Effluent
                                                                        PH
10 percent  of the total number  of samples.   As  an independent  check,  a
U.TJ.  EPA reference sample should  be analyzed at  a minimum of every 3 mo.

     Corrective  Action—If the  pH meter does not  appear to be operating
correctly, consult the manufacturer's  troubleshooting guide.   Common  problems
include  a dirty  electrode,  failure to  fill  the  reference portion of  the
electrode with internal solution,  and  inadequate stirring.

     Data Quality and Reporting—A precision of  +_ 0.02 pH unit  and an accuracy
of 0.05 pH unit can be  achieved   under  the best circumstances.  However,
the '.imit of accuracy under most  circumstances  is ^0.1 pH unit (APHA 1985).
A precision  of 0.1 pH unit is considered acceptable  (U.S. EPA 1979b).

     Measurements of pK are reliable only when  the instrument  has been
calibrated by standard buffers bracketing the-desired  range.  Samples having
a pH greater than  10 may require  a special probe to  correct for  interference
from sodium ions.

     It  Is  recommended that pH values be reported  to the nearest 0-. 1 unit.
In addition, the  ambient  temperature  at the time  of measurement  of each
sample should be reported  to the nearest  degree C.   The  results of all
determinations should  be reported,  including QA replicates.  Any  factors
that may have influenced sample quality should  also  be reported.
                                     27

-------
                                                                  Effluent
                                                               Temperature
Temperature

Field Procedures—

     Col lection—Temperature can be measured using a mercury-filled Celsius
thermometer on samples  collected in glass or plastic containers.  The  ther-
mometer should have a  scale etched .jn capillary glass for O.io C increments
and a minimal thermal capacity to permit rapid equilibration.  Temperature
can  be  measured in  situ using a reversing thermome >r  or a thermistor.
Of these two  in  situ instruments,  the thermistor is more accurate, but
also more expensive.

     Processing—Because  temperature can  change rapidly  after  a sample
is removed from ambient  conditions, temperature determinations by thermometer
should be made immediately  after sample collection.

Laboratory Procedures—

     Analytical  Procedures—Methods  for  making temperature measurements
are described in U.S.  EPA Method 170.1 and in APHA Method 212.  It is critical
that the measuring  device be adequately immersed in the  sample and allowed
to completely equilibrate (i.e., the temperature reading stabilizes)  before
temperature is read.

     Calibration and Preventive Maintenance—Each kind  of temperature-measuring
instrument should be calibrated  frequently  against a National Bureau of
Standards (NBS)-certified thermometer that  is  used with  its certificate
and correction chart.   An NBS thermometer  is recommended because  some commercial
thermometers may be as  much as  3°  C  in  error  (APHA  1985).

     To  prevent breakage, it  is  recommended  that each  thermometer be enclosed
in a metal case.  If a  mercury  thermometer  is  broken, samples  or bottles
in the  vicinity of the  exposed  area  may be  contaminated by the  mercury.

                                      28

-------
                                                                  Effluent
                                                               Temperature

     Quality  Control  Checks—Each temperature-measuring instrument should
be calibrated against an NBS thermometer at least every week.   It  is recommended
that  calibration should be conducted  daily when  a  temperature violation
is suspected.

     Corrective  Action—If  the temperature-measuring  instrument cannot
be calibrated consistently  against the NBS thermometer,  it  should be repaired
or replaced.

     Data Quality and  Reporting—Precision and accuracy have not been determined
for temperature measurements  (U.S.  EPA 1979b).   Temperature measurements
should be reported to  the  nearest O.io C.  The results  of all  determinations
should be reported,  including QA replicates  and  calibration checks.   Any
factors that may  have  influenced  sample quality should  also be reported.
                                      29

-------
                                                                 Effluent
                                                                Turbidity

Turbidity

Field Procedures--

     Col lection--Turbidity samples  can  be collected  in  glass or plastic
containers.   Samples should be removed from  the sampler as soon as possible
after  collection to minimize settling of suspended material within the
sample''.  Sample containers and lids  should  be rinsed thoroughly with sample
water before samples are collected.

     Process ing--Because turbidity  samples cannot be preserved adequately,
they should  be analyzed as soon as possible  after collection.  If  a  delay
occurs, samples should  be  held at  40 c for no more than 48 h to minimize
microbiological  decomposition of solids.   The  length of  delay should be
noted on the log sheet.

Laboratory Procedures —

     Analytical Procedures—Two procedures are available for  turbidity
measurements in effluent:  the nephelometric method and  the  visual  method
(APHA  1985).  Although  the results of both methods are comparable  in a
general manner (U.S. EPA  1979b), they are  not related directly  (AHPA 1985)
and  therefore  should  not be used interchangeably.  Because of its greater
precision, sensitivity, and  applicability  over a wide turbidity range, the
nephelometric method  is generally preferable to the v'isual method (APHA 1985).

     The  nephelometric method  is based on  the nephelometric turbidity  unit
(NTU),  and is described  in U.S. EPA Method 180.1 and  in  APHA Method 214A.
For  turbidities greater than 40 NTU, samples  should  be diluted  with one
or more volumes of turbidity-free water  until  the turbidity  falls below
40 NTU.
                                      30

-------
                                                                 Effluent
                                                                Turbidity

     The visual method  of  turbidity determination  is based on the Jackson
turbidity unit  (JTU), and is described in APHA Method  214B.   It currently
is the  method  specified by  the California Ocean  Plan.  The lowest turbidity
that can be measured using  a candle  turbidimeter is  25 JTU (APHA  1985).
Samples  with turbidity values less than 25 JTU must  be estimated indirectly
by visual comparison with standards.  Samples having turbidity values exceeding
1,000 JTU  should  be diluted with one  or more volumes of  turbidity-free
water until turbidity falls  below  1,000 JTU.

     Turbidity-free water  is  distilled  wa:er  passed  through a  membrane
filter having a pore size of 0.2 urn.  Prior to analysis,  each sample  should
be shaken  well to thoroughly disperse solids and resulting air bubbles
should be allowed  to dissipate before the sample  is analyzed.

     Interference with  turbidity measurements arises from several sources.
The presence of floating debris  and coarse sediments  that settle out  rapidly
will  give low readings.   Therefore,  turbidity readings should be made as
soon as possible after sample  agitation.  Finely divided air bubbles  will
affect  results in a positive  direction.   If  present, the bubbles should
be allowed to  dissipate before taking the  reading.   Finally, dissolved
substances in  the  sample, which  absorb light, will reduce turbidity readings.
Any color  in  the sample  should therefore be noted.

     Calibration and Preventive Maintenance—It  is recommended  that standard
suspension of  formazin  be used  to calibrate the nephelometer.   Formazin
provides a more  reproducible  turbidity standard  than do  other materials
used  in the past.   The  formazin  standard  suspension  should  be  prepared
daily (APHA 1985).  Commercially available standards, such as styrene  divinyl-
benzene beads  (trade name AMCO-AEPA-1),  can be substituted  for  formazin
if they are demonstrated  to  be  equivalent to freshly prepared formazin
(APHA 1985).  Standards  measured on the nephelometer should cover the range
expected  for  the samples.  At least  one  standard should be run  in  each
 instrument range to be  used.  The instrument should  give  stable readings

                                     31

-------
                                                                  Effluent
                                                                 Turbidity

in  all  sensitivity ranges used.   If a  precalibrated scale is not  supplied,
calibration curves  should be prepared for each  instrument range to  be used.

     Standards  for visual comparisons  based  on JTU should be prepared  from
natural turbid water or kaolin, and should be  calibrated weekly using the
candle turbidimeter.  The  candle used  in the turbidimeter should be  made
of beeswax and spermaceti, and should  burn  within the  limits of 114-i26
grains/h.

     Quality  Control  Checks—The nepheloneter should be calibrated at the
start of each series  of  analyses  and  after  each group  of 10 successive
samples.  Duplicate  analyses should  be conducted on at least 10 percent
of the total number of samples, using either method of turbidity determination.

     Corrective Action—If the  nephelometer wiM not  stabilize in any of
the relevant ranges or if the  instrument does  not  appear  to be functioning
properly  in  any other aspect, the  manufacturer's troubleshooting  guide
should  be  consulted.   Sample tubes that become scratched or etched should
be  replaced.

      Data  Quality and  Reporting—Limited  precision data  for  the nephelometric
method indicate that  standard deviations of measurements vary  directly
with the level  of turbidity  (U.S.  EPA 1979b).   Accuracy  data are not available
at  present.   The sensitivity  of  the nephelometer  should allow  detection
of a  turbidity difference of  0.02 unit or less in waters having turbidities
 less than  1.0 unit.  Results should be  reported in nephelometric  turbidity
 units (NTU).  U.S. EPA Method  180.1 and APHA Method 214A describe  the nearest
 reporting  units as a function  of  the range of values measured.

      Results of analyses using  visual comparisons should be reported  in
 Jackson turbidity units (JTU).   APHA Method 214B describes  the nearest
 reporting units as a function of the  range  of values measured.   The method
 used to determine each JTU  value (i.e., candle turbidimeter or bottle  standards)
 should be identified.
                                       32

-------
                                                                  Effluent
                                                                 Turbidity


     Results  of  all  turbidity determinations  should be reported, including
QA replicates. Any factors that may have  influenced sample  quality  should

also be reported.
                                      33

-------
                                                                   Effluent
                                                     Total  Suspended  Solids
Total Suspended Solids

Field Procedures--

     Col lection--Samples  should be collected  in glass or plastic  bottles.
Nonrepresentative particulates,  such as leaves, sticks,  or  rocks, should
be noted and excluded from  the  sample analysis.

     Processing—Total suspended  solids  samples  cannot be preserved adequc. ?ly
and  should therefore be analyzed  as  soon as  possible aftei col lection (A.'HA
1985).   If a delay occurs,  samples  should  be held at 40 C to minimize micro-
biological decomposition of solids.  The length  of delay  should not  exceed
7 days and should be noted  on the log sheet.

Laboratory Procedures—

     Analytical  Procedures — Suspended solids  determination should be made
according  to procedures described in U.S.  EPA Method  160.2 or APHA  Method
209C.   The drying temperature of the  filtered residue  can  influence results
because temperature and length of heating  affect weight  losses due to volatili-
zation  of  organic matter, mechanically  occluded water, water of crystallization,
and  gasses  from  heat-induced  chemical decomposition,  as well as  weight
gains due  to oxidation (APHA 1985).  Thus, drying temperature must be carefully
controlled and not allowed to deviate  from the recommended range of 103-1050 C.

      To avoid contamination, filters  should be handled using forceps during
 all  steps  from  initial  to final  weight determinations.   Filters should
be washed  with distilled water and dried prior to initial  weight determination.
 Because glass-fiber filters are hygroscopic, they must be stored  in a desiccator
when cooling.  When  filtering the samples, it is critical  that  the filter
 is seated  tightly on the surface of the filtration  apparatus  and  that all
 holes  in  the crucible  are covered.   To ensure complete removal  of  salts
 after filtering  the sample, the  filter should  be rinsed  with  a minimum
 of  three  successive 20-mL portions of distilled water.  It is recommended
                                      34

-------
                                                                 Effluent
                                                    Total  Suspended Solids

that  enough water be filtered to  ensure that at least  5  mg of residue Is
collected.  Because excessive residue on the filter may form  a  water-entrapping
crust,  the sample size should be  limited to that  which  yields less than
200 mg of residue  (APHA 1985).  Prolonged filtration  times resulting  from
filter  clogging may produce high  results due to'excessive solids capture
on the clogged  filter.  Therefore,  filtering should  be terminated  before
any evidence of clogging is noted.

     jlaMbration  and Preventive Maintenance—The analytical balance should
be calib ated weekly  using standard weights, according to  the manufacturer's
instructions.   It is recommended that the balance have  a  minimum accuracy
of 0.1 mg.  The manufacturer's preventive maintenance procedures  should
be followed carefully.

     Quality Control Checks--For each weight determination, filters should
be run repeatedly  through the drying/cooling cycle until the weight  loss
is  less than 4 percent of the previous weight or 0.5 mg, whichever  is less
(APHA 1985). Duplicate analyses should be conducted on at least  10  percent
of the total number of samples.  U.S. EPA reference samples  should be analyzed
quarterly to check the overall accuracy of the method.

     Corrective Action—If the analytical balance will not produce repeatable
measurements at 0.1 mg, the manufacturer's  troubleshooting guide  should
be  consulted.   If the filter becomes clogged during filtration,  it should
be discarded and  the analysis should be repeated using  a clean  filter.
To  prevent clogging  of  the second filter, the volume of sample analyzed
should be reduced.

     A  filter  blank should be carried  through  the preparation, drying,
and desiccation steps to  monitor  any changes  in  filter tare  weight.   If
the  filter weight is not consistent after drying  and cooling,  the  desiccant
should be checked.  A color-indicating desiccant  is recommended, so that
spent desiccant is easily detected.   Also, the  seal on the desiccator should
be  checked  and, if necessary, ground glass  rims should  be greased or  "0"
                                     35

-------
                                                                 Effluent
                                                    Total Suspended  Solids

rings  replaced.  Cooling  should  be  closely timed,  so  that weighing  times
are consistent between  batches  of samples, thereby minimizing the need
for multiple weighings.

     Data Quality  and Reporting—Precision of results varies directly with
the concentration of  suspended  matter and, at  low  levels, the ratio of
the weight  of the suspended matter  to the weight  of  the  filter.   There
are no procedures  for  determining  the accuracy  of field measurements of
suspended matter.  Total  suspended  solids measure'.  Tits should be reported
as mg/L to a minimum of two significant figures.  Result., of all determinations
should  be  reported,  including  QA replicates.  Any factors that may have
influenced  sample quality should  also  be reported.
                                     36

-------
                                                                 Effluent
                                                         Settleable Solids
Settleable Solids

Field Procedures—

     Collection—Settleable solids  samples should be collected  in  1-L glass
or plastic containers.  The  sample  container  selected for  use  should be
checked *.o ensure that material  in suspension does not adhere  to the container
walls.  Nonrepresentative particulates, such as leaves, sticks,  or  rocks,
should be noted and excluded from  sample analysis.

     Processing—Settleable solids  samples cannot be preserved  adequately
and should therefore be analyzed as  soon as possible after collection (APHA
1985).  If a delay occurs, samples should  be held at 40 C to minimize micro-
biological decomposition of solids.   The length of delay should not exceed
48 h and should be noted on the log  sheet.

Laboratory Procedures—

     Analytical  Procedures—Procedures used  to  determine settleable solids
concentrations are presented in U.S.  EPA  Method 160.5  and  in APHA Method
209E.  The sample should  be  well-mixed  before introduction to the Imhoff
cone.   It  is critical  that  floating material  is not  included with settleable
material.

      Calibration  and  Preventive Maintenance—Not applicable.

      Quality Control Checks—Duplicate  analyses  should be  conducted on
 at least 10 percent  of the  total number of samples.

      Corrective Actions—If  suspended  material is found  to  adhere to the
 sides of the sample collection containers, a  different type  of  container
 should be used.
                                      37

-------
                                                                   Effluent
                                                           Settleable Solids

     Data  Quality  and  Reporting--The practical  lower  limit of measurement
defends on sample composition and generally is in the range  of 0.1 to 1.0 mL/L
(APHA  1985).   Precision and  accuracy  data are not available at present.
Settleable solids concentrations should  be reported  as mL/L to  a minimum
of two  significant  figures.  Results should be reported for all  determinations,
including QA replicates.   Any factors that may have influenced sample quality
sr.ould  also be  reported.
                                        38

-------
                                                                 Effluent
                                                     Floating Particulates

Floating  Particulates

Field Procedures--

     Col 1ection--A minimum of 5 L of sample should be collected  in a glass
or plastic container.  The container should be rinsed thoroughly with sample
water before sample collection, as no preservation techniques  are available
for this  variable.

     Processing—For  comparable results, samp  >s must be treated uniformly
throughout sampling and  handling.   Analysis  should be  performed as  soon
as possible after sample  collection, as no  preservation techniques are
available for this variable.

Laboratory Procedures--

     Analytical  Procedures--Floating  particulates  should be analyzed  in
accordance with APHA Method 206A.   At  present, this method  is tentatively
recommended by APHA.   Because even  slight differences 'in  sampling and handling
can give  large differences  in measureable floating  particulates, all  samples
should  be treated uniformly,  preferably by adequately mixing them before
flotation.  When mixing  the sample, care should  be  taken  to avoid extensive
air  entrapment  through formation of  a large vortex.   Because temperature
variations can affect  results, all  tests should  be  conducted at a constant
temperature, preferably  200 c.

      Calibration and Preventive  Maintenance—No calibration  procedures
apply to  the recommended method.   To prevent  oil and  grease from sticking
to the analytical  equipment,  all internal  surfaces should  be coated with
TFE.

      Quality Control Checks — It  is  recommended  that  duplicate analyses
be conducted on  a minimum of 5 'percent of  the total number of samples,
with an  additional 5 percent of the samples checked for recovery.
                                     39

-------
                                                                   Effluent
                                                      Floating Particulates
     Corrective Actions--If oil  and  grease  appear  to be sticking to surfaces,
7FE coatings should be renewed.   If  recovery  drops below 90 percent, samples
snould be analyzed again and each step of the analysis carefully scrutinized.

     Data  Quality and  Reporting--The minimum detectable concentration of
f'oating particulates using the  recommended  method is  1  mg/L.  Precision
varies with the concentration cf suspended  matter  in the sample.  A coefficient
of variation >  ' 5.7 percent has  been achieved using  five  replicate samples
(APHA  1985).  Although there is no completely satisfactory procedure for
determining the  accuracy of the method, approximate  recovery can be determined
by running a  second test for floatables on  all water discharged throughout
the analytical  procedures, except for the last 10 ml.  Typical recoveries
exceed 90 percent (APHA  1985).   Concentrations of floating particulates
should be reported in mg/L to the nearest 0.1 unit.  Results for all deter-
minations should  be reported,  including QA  replicates and recovery checks.
Any factors that may have influenced sample quality  should also be reported.
                                       40

-------
                                                                 Effluent
                                                Dissolved Oxygen (Winkler)
Dissolved Oxygen  (Winkler Method)
Field Procedures—
     Col lection—Prior  to  sample collection,  the fixing reagents should
oe prepared and the dispensing apparatus should be  filled.  The accuracy
of the volumes being dispensed should be checked  and no  air should be trapped
in the system.   It is recommended  that 300-mL glass  800 bottles  having
ground  glass  stoppers be used for  the Winkler method.  Becaus-  chlorine
interferes with the Uinkler method, samples should  be taken a   a  point
prior to chlorination.

     Grab  samples  for dissolved  oxygen should  be  collected very carefully
to avoid aeration and they should be  preserved as soon  as  collected.  Composite
sampling  for  dissolved  oxygen is  not possible.   If  possible, sample from
a tap by attaching a soft-walled rubber tube  to  the  tap and extend  the
tube  to the bottom of the sample  bottle.   The tubing should be flushed
with sample water to remove  air bubbles, and the  bottle  and stopper  should
be rinsed at  least three times  with  sample water.   The bottle should be
filled slowly until at least half  full, and then  filled  rapidly.  At  least
one  full  bottle  volume of sample  should overflow the bottle before the
tubing is removed, slowly.   After  the tubing is removed,  the stopper  should
be carefully  put in place with  a twisting motion  while water  is displaced
from the bottle.   Once stoppered,  the sample  should  be checked  for  air
bubbles.   If  bubbles are present,  the sample should  be discarded and a
new sample collected.  Fixing solutions should be  added to  each  sample
immediately after collection.

     Processing—The  stopper  should  be  carefully removed from the bottle
without agitating the sample.  Each  fixing reagent  should be added by gently
placing the tip  of the  pipet slightly below the  surface  of the sample and
gently pushing the plunger.  The plunger should  not be released until  the
pipet  has been  removed  from the sample.  The pipet  tip should be rinsed
with distilled water before  being  returned to the reagent bottle.
                                     41

-------
                                                                  Effluent
                                                 Dissolved  Oxygen  (Uinkler)
     After the fixing reagents  have been added, the bottle should be carefully
stoppered without introducing air bubbles.  Excess fluid around  the  outside
of the stopper should be poured  off and the sample bottle should be  inverted
5-10 times to thoroughly disperse  the precipitate.  It is also  a good  practice
to  invert bottles  several times approximately  20 min after  fixation, to
ensure thorough dispersion  of the precipitate.

     After allowing the precipitate to settle  for 10-15 minutes, the  :»< ipper
should be removed and sulfuric  acid  should be added ti the sample i.i the
same  manner as  the  fixing reagents.   The stopper should then be  replaced
and the bottle Inverted until  all  of the  precipitate  has  dissolved.  If
the  precipitate  fails to dissolve, it should  be allowed  to  settle  again
and additional sulfuric acid should be added to the sample.  It  is  critical
that  all  of the  precipitate be dissolved  before samples are stored.   Also,
it is critical that samples not be allowed  to  stand longer than 8  h before
sulfuric acid is added, as  erroneous  measurements may result.

     Preserved dissolved  oxygen samples  should  be  stored in the  dark at
10-20° C.  Samples should  be analyzed as soon as possible after collection,
and storage time should not exceed 8 h.  The  length of storage should be
recorded on the log sheet.

Laboratory Procedures--

     Analytical Procedures—The modified Winkler  method is described  in
detail  in U.S. EPA Method 360.2 and  in APHA Method  421B.

      Calibration  and Preventive Maintenance—Methods  of standardizing the
thiosulfate  solution  are presented  by U.S.  EPA (1979b).   It  is  recommended
that  one  person  perform the standard  and  sample titrations because of subjec-
tivity  in the color of the endpoint.
                                      42

-------
                                                                  Effluent
                                                Dissolved Oxygen (Ulnkler)

     Preventive maintenance is limited  to  ensuring that reagent dispensing
and titrating  equipment is clean and  functions properly.

     Quality  Control  Checks—All  standard  titrations should be dupl icated.
It is recommended that duplicate analyses  be  conducted on at least 10 percent
of all  samples.  Reagent blanks should  be run whenever a reagent is changed.

     Corrective Action—If the results  obtained by  running duplicate standard
titration  of  the thiosulfate solution do not  agree within HJ.05 tnL,  the
titratior-j should  be repeated until  agreement  is achieved.  All reagent
dispensers should be checked for bubbles,  and the amounts of reagents delivered
should  be verified.

     Data Quality and Reporting—Using  the modified Winkler method, reproduc-
ibility for field samples  is approximately 0.2 mg/L  of dissolved  oxygen
at  the  7.5 mg/L  level  (U.S.  EPA 1979b).  Duplicate  titrations made during
standardization of reagents should  agree within  ^0.05 ml.  With careful
collection and  treatment  of samples, dissolved oxygen  as  low as 1 percent
of  saturation can be  measured.  Dissolved oxygen concentrations  should
be  reported  in  mg/L to  the nearest  0.1 unit.   Results should be reported
for all determinations,  Including QA replicates  and  reagent blanks.   Any
factors that may have  influenced sample quality  should  also  be reported.
                                      43

-------
                                                                 Effluent
                                                  Dissolved  Oxygen (Probe)
Dissolved Oxygen  (Probe Method)

Field Procedures--

     Col 1ection--Grab  samples for dissolved oxygen should  be collected
very carefully  to avoid aeration.   Composite sampling for dissolved  oxygen
is not  possible.   If possible,  sample  from a  tap by attaching  a  soft-walled
rubber tube to the tap and  extend the tube  to  *• -e bottom of  the  sample
bottle.  The tubing should be flushed with sample water to remove air bubbles,
and the bottle  and  stopper rinsed  at least three  times with sample  water.
The bottle should be filled slowly until at least half full,  and  then filled
rapidly.  At least  one full  bottle volume of sample should  overflow the
bottle  before the tubing  is  removed  slowly.  After the tubing  is removed,
the stopper should  be carefully put in  place with a twisting  motion while
water  is  displaced from  the  bottle.   Once  stoppered, the  sample should
be checked  for air bubbles.   If  bubbles are present, the sample  should
be discarded and  a  new sample collected.

     Processing—Because no reagents are used  to preserve the oxygen samples,
analyses should be conducted  immediately after collection.   If  a delay
occurs, it should be noted on the  log sheet.

Laboratory Procedures—

     Analytical Procedures—Detailed  analytical procedures should be  provided
by the manufacturer  of  the dissolved oxygen meter.   General  procedures
are  listed  in U.S.  EPA Method 360.1 and APHA Method  421F.

      jeveral  precautions should be  taken when making measurements with
a membrane  electrode.   First, constant  turbulence  should  be provided by
a  stirrer  to  ensure precise  measurements.   Second,  adequate time should
be allowed  for the  instrument to warm up  before measurements  are started,
and, as individual  samples are  analyzed,  the probe  should be allowed to
                                      44

-------
                                                                  Effluent
                                                  Dissolved Oxygen (Prooe)
stabilize  to sample temperature  and dissolved  oxygen.  Third,  reactive
gases, such as  chlorine  and  hydrogen sulfide,  pass through  the membrane
probes and  may  interfere with the analysis or desensitize the probe.  Finally,
broad variations  in the kinds and  concentrations of salts  in samples can
influence  the  partial  pressure of oxygen  in  samples and thereby  affect
measurement accuracy.

     Calibration and Preventive Maintenance—Cal ibration procedures  should
follow the instructions given .y the manufacturer of the  dissolved oxygen
meter.   The meter generally can  be calibrated using one of three methods:
Winkler titration,  saturated water,  or  air.   The air  method is simplest
and quite reliable.   Overall  error  is diminished when the probe and instrument
are calibrated  under  conditions  of  temperature and dissolved oxygen that
match those of  the  samples.  Calibration  can  be disturbed by physical  shock,
touching the membrane, or desiccation of  the  electrolyte.

      Preventive maintenance procedures should  follow the manufacturer's
reccmnendations.  The oxygen probe should always  be  stored in a humid environ-
ment  to prevent drying out and the need to frequently replace membranes.

      Quality Control Checks—The instrument  should   be calibrated  at the
beginning of each series of measurements and after  each group of 10 successive
samples.  Duplicate measurements  should  be made on  at least 10 percent
of the total number of samples.

      Corrective Action—If  the dissolved  oxygen meter does not appear to
be operating  correctly, consult the manufacturer's troubleshooting guidelines
 for remedial actions.

      Data  Quality and Reporting—Repeatability of dissolved oxygen measurements
 using a membrane electrode should  be 0.1 mg/L  and accuracy  should  be *_ 1
percent (U.S.  EPA 1979b).   Sensitivity of the  electronic  readout meter
 for the output from the  dissolved  oxygen probes should  normally  be  0.05 mg/L
                                      45

-------
                                                                  Effluent
                                                  Dissolved Oxygen  (Probe)
(U.S.  EHA 1979b).  Dissolved  oxygen concentrations should be  reported in
mg/L to the nearest  0.1  umt.  Results should be reported  for  all  deter-
minations, including QA  replicates. Any  factors that may have influenced
sample quality should also be reported.
                                      46

-------
                                                                  Effluent
                                                                       BOO
Biochemical  Oxygen  Demand (BOD)

Field Procedures--

     Collect ion--BOD samples can be collected  in  glass or plastic containers.
Sample containers and caps should be rinsed thoroughly  with sample  water
before sample collection.

     Processing--BOD samples should be analyzed • mediately after collection.
If a delay occurs, samples should he refrigerated at 40 C  to minimize reduction
of BOD.  Samples should  not be stored  for more  than  48 h and the  length
of storage should be recorded on the log sheet.  Refrigerated  samples  should
be warmed to 200 C prior  to analysis.

Laboratory Procedures--

     Analytical Procedures—BOD concentrations should be  determined according
to U.S. EPA Method 405.1  or APHA Method 507.   Samples having more oxygen-
demanding materials than the amount of oxygen in air-saturated  water  should
be diluted to  balance  the oxygen demand and supply.  If samples  are diluted,
nutrient  addition  (I.e.,  nitrogen,  phosphorus, trace metals)  and pH buffering
of the dilution water are necessary to ensure  that the  sample  is  suitable
for  bacterial growth.   To prevent air from  infiltrating  the  incubation
bottles,  a water seal should be used.  When  samples are  incubating,  all
 light  should  be excluded to  prevent photosynthetic production of oxygen.
 Samples containing caustic  alkalinity or acidity should  be  neutralized
 to pH  6.5-7.5 usfng sulfuric  acid or  sodium  hydroxide.  Samples containing
 residual  chlorine  must be dechlorinated  (e.g.,  using sodium thiosulfate).

      Calibration and Preventive Maintenance—Dissolved oxygen concentrations
 should be measured on all  dilution water blanks and  seed controls.   The
 dissolved  oxygen uptake of  the  dilution water should  not exceed 0.2 mg/L.
 The dissolved oxygen uptake of seeded  dilution water should be between
 0.6 and  1.0 mg/L.  A glucose-glutamic acid  standard check  solution should
                                      47

-------
                                                                Effluent
                                                                     BOO
be  incubated  with each batch of samples.   Dissolved o.tygen measurements
should be calibrated according  to accepted procedures (e.g., see descriptions
of the Winkler and probe methods in this document).

     Quality Control  Checks—The dilution water blank provides a quality
control  on the dilution water  as well as on the cleanliness  of  analytical
equipment (e.g.,  incubation bottles).   Each sample should  be  analyzed at
a -ninimum of three different dilutions to ensure that  dissolved  oxygen
uptake  is  i  the  optimal range.  Optimal  dissolved oxygen uptake is at
least 2 mg/l  after the  incubation,  with a residual dissolved oxygen of
at  least 1 mg/l in the sample.  Duplicate analyses should be conducted
on at least 10 percent of the  total number of samples.

     APHA  (1985)  should be  consulted  for  methods  of correcting for the
many kinds of  interference that may accompany BOD analyses.

     Corrective  Action—If the dilution water blanks exceed 0.2 mg/L,  clean-
liness of containers and water should be checked.  Containers  may  require
1+1 HC1  rinse  after detergent washing to remove any residual organic material.
Containers rinsed with  acid  should be thoroughly rinsed with distilled
water to prevent any acid carryover.

      If  a  2-percent  dilution of the  glucose-glutamic acid standard check
solution is outside the  range of 200 +_ 37  mg/L,  BOD determinations made
with the  seed and dilution  water  should  be rejected.   The problem, which
could arise from numerous sources, may  be  identified by running a  series
of dilution water blanks using  different water sources with and without
seed, preparing  a  fresh  solution of glucose-glutamic acid, changing the
seed, or  preparing fresh reagents for  the  dilution water.  The source  of
the  problem needs  to be  determined before  performing any additional BOD
analyses.

      Data Quality and Reporting — Precision  data for spiked natural waters
 indicate that standard  deviations of  +_0.7 and +26 mg/L can  be achieved
                                     48

-------
                                                                  Effluent
                                                                      BOO
for  BOD  concentrations of 2.1  and  175 mg/L, respectively (U.S.  EPA  1979b).
There is no acceptable method for determining the accuracy of the BOO test.
BOD  data should be reported as mg/L to  the nearest 0.1 unit.   Results for
all determinations should be reported, including QA replicates,  dilution
*ater  blanks, and glucose-glutamic acid  standards.  Any factors that may
have influenced  sample quality should also be reported.
                                      49

-------
                                                                  Effluent
                                                   Total Chlorine Residual
Total Chlorine Residual

Field Procedures--

     Col lection—Chlorine residual  samples  can  be  collected in glass or
plastic containers.   During collection,  samples should  be  protected from
strong light and excessive agitation.

     Process ing--Because chlorine in aqueous solution  is  not stable.  cMorine
content of samples  will  decrease rapidly after collection,  part cularly
for  weak solutions.   In  addition, both strong  light (e.g., sunlight)  and
agitation will  accelerate  the reduction of chlorine.   There is no preservation
technique available.   Chlorine-residual samples should therefore be analyzed
immediately after collection and protected from strong  light and  agitation
before and during the analysis.

Laboratory Procedures--

     Analytical Procedures--Chlorine residual  should  be determined using
U.S. EPA Methods 330.1, 330.2, 330.3, 330.4, and 330.5  or APHA  Methods
408A,  408B, 408C, 4080,  and 408E.   The methods should be consulted to determine
which are most appropriate for available equipment,  expected effluent constit-
uents,  and expected levels of precision and accuracy.   The amperometric
method (i.e.,  U.S. EPA Method  330.1,  APHA Method 408C) is the  preferred
method  because it Is not subject to interference  from  color,  turbidity,
 iron,  manganese, or nitrite nitrogen (APHA  1985).  This method  is not as
 simple  as  the colorimetric techniques  and requires greater  operator skill
 to obtain the best results.  However, sample  color and turbidity may ir'erfere
 with all  colorimetric procedures.  Because all methods of  chlorine-residual
 determination depend on the  stoichiometric  production of iodine, effluent
 containing  iodine-reducing substances may  not  be analyzed accurately by
 any  of these  methods, especially  where iodine  remains in solution.
                                      50

-------
                                                                 Effluent
                                                   Total  Chlorine Residual
     Calibration and Preventive Maintenance—The methods should be consulted
for procedures related to preparation of standards and cal'oration.

     Quality Control  Checks — For all colorimetric procedures, a minimum
of one color  and one turbidity blank  should be  run  per batch to evaluate
these  potential  sources of interference.  For  all  techniques, duplicate
analyses should be conducted on a minimum of 10 percent of  the total  njmber
of samples.

     Corrective Action—There are many sources of •' terference with chlorine
residual tests.  If a particular source of  interference cannot be adequately
compensated for using one method, an alternate method should  be used.

     Data Quality and  Reporting—The detection and accurate quantification
of total chlorine residual in effluent is  routinely attainable, although
method  detection  limits, accuracy,  and precision can vary depending on
the method  used.  The method should  therefore  be consulted to determine
expected detection limits, accuracy, and precision.  Total  chlorine residual
concentrations should be reported as mg/L to the nearest 0.01 unit.  Results
should  be  reported for all determinations,  including QA replicates and
color and turbidity blanks.  Any  factors that  may have influenced  sample
quality should also be reported.
                                     51

-------
                                                                 Effluent
                                                            Oil  and Grease
Oil and Grease

Field Procedures—

     Collection--Samples should  be  collected in glass  bottles.   Bottles
should first be  washed  with  a warm aqueous  detergent mixture,  and then,
in  sequence, thoroughly  rinsed with hot  tap water, rinsed at  least twice
with distilled water,  rinsed with 1,1,2-trichloro-l,2,2-trifluoroethane
(i.e.,  Freon or equivalent), and dried  in  a  -.lean oven at >_105o  C for 30
min.  Bottle caps should be lined with TFE-coated cardboard inserts  or  aluminum
foil.  Plastic containers  are not acceptable.

     Only  grab  samples,  and  not  composite samples,  should be collected
to  limit loss of oil  and grease on  sampling equipment.   The entire grab
sample should  be submitted  for analysis.  To obtain a daily average concen-
tration, grab samples  collected at prescribed time  intervals can be submitted
for analysis.  Headspace should be left in each  sample container for addition
of  acid and mixing.

      Processing—Acidify the sample in the collection bottle to a  pH  <2
using  sulfuric or hydrochloric acid.  Samples should be  stored in the dark
at  40  c.   Recommended maximum holding time is 28 days  (U.S. EPA 1984).

Laboratory  Procedures—

      Analytical  Procedures—Oil  and grease should  be  analyzed using procedures
described  in U.S. EPA Method 413.1 or APHA Method  503A.   Because asphaltic
materials  are  insoluble in  Freon,  the recommended  method will  give low
 recoveries for  samples  containing such material.   U.S. EPA Method  413.1
 (gravimetric  oil  and grease) measures  relatively nonvolatile hydrocarbons,
 vegetable oils, animal  fats, waxes, and  soaps.   The method  is  often  used
 for wastewater analyses  since it requires minimal instrumentation  and cali-
 bration.   Sulfur causes interference for  the  gravimetric method because
 it  will  be extracted  and  included as oil  and grease.  Light  hydrocarbons
                                      52

-------
                                                                 Effluent
                                                           Oil and Grease

that volatilize  at temperatures below 700 c (e.g., gasoline through No.2
fuel  oil)  are  lost during  the  solvent removal step.

     Calibration and Preventive Maintenance—For gravimetric oil and grease
analyses,  check accuracy of  the analytical balance  periodically  (minimum
of once per  week recommended) using  Class S weights.   A service contract
that includes  scheduled preventive maintenance  at  least once per  year  is
recommended.   Scratched, chipped, or cracked boiling  flasks should  be replaced.
For infrared  oil  and grease  analys.', follow the manufacturer's  preventive
maintenance  procedures  for  the  i.ifrared  spectrophotometer.  Cells used
for analysis should be  checked  for  scratches  each  time they are  used.
Scratched  cells should not be used.

    .Quality  Control  Checks—Duplicate samples should be collected and
performed  at a minimum of  every  10 samples  to establish  an estimate of
precision.  Because samples should not be  split after collection,  separate
grab samples  should be taken for  analysis.

     Distilled water spiked with a U.S. EPA performance sample  should be
extracted and analyzed  with every batch of samples to  monitor recovery.
Bottles should  be checked for  cleanliness  by analyzing  distilled water
that has been acidified  in a sample bottle.   A  solvent blank should  accompany
each batch of samples.

     Corrective  Action—If  oil  and  grease  concentrations  in procedural
blanks  are greater than the detection limit, check  the cleanliness  of  all
glassware.  Always use separatory funnels with TFE  stopcocks to  avoid contam-
 ination from  stopcock grease.  For gravimetric analyses, high results will
be obtained  if any Freon  or fumes remain in the flask  after  distillation.

      If difficulty occurs with emulsions during solvent extraction,  follow
 procedures described in the methods.  If  the  emulsion still fails  to  dissipate
                                      53

-------
                                                                  Effluent
                                                            Oil and Grease

after addition of salts,  gently turn the separatory funnel  to the horizontal
position and slowly rotate.   Be  careful to keep  the  stopper securely in
place.

     If precision or recovery  from spike results are poor, check adequacy
of extraction by increasing shaking time.

     Data Quality and Reporting—The definition of oil  and grease  is based
on the procedures  used,    u  less identical procedures are used,  oil  and
grease  deter,.,inat-ions are ,iot intercomparable.  Therefore, the method used
for analysis should always be specified.

     Inter laboratory recovery  results were 102+37 percent at 6.0 mg/L and
97+35 percent at 18.0 mg/L  for the  gravimetric method (U.S. EPA  1983a).
Measurements should be  reported  to  a minimum of two significant  figures
in mg/L.  Detection limits are in the range of  5 mg/L for gravimetric oil
and grease determinations (U.S. EPA Method 413.1).   Results should be reported
for all determinations,  including QA replicates, spiked samples, and  blanks.
Any factors that may have Influenced sample quality should  also be reported.
                                       54

-------
                                                                  Effluent
                                                        Nitrogen (Ammonia)

Nitrogen (Ammonia)

Field Procedures--

     Conection--Ammonia samples can be collected  in glass or plastic bottles.
Each bottle and  cap should  be rinsed  thoroughly with  sample water  prior
to sample collection.

     Processing--Resu1 ts of ammonia analyses  are most   ^liable when they
are made on fresh  samples.  However, if analysis must  be  relayed,  samples
can  be  stored for up to 28  days by  acidification to pH<2 with sulfuric
acid  and refrigeration at 40 c.  The length of delay before  analysis should
be recorded on the log sheet.

Laboratory Procedures--

     Analytical  Procedures—Manual distillation of effluent  samples prior
to ammonia determinations is required by the U.S.  EPA,  unless  data  on file
demonstrate that distillation  is not required.  Distillation is  recommended
due to the sensitivity of ammonia procedures to color and possible  interferences
in  the  effluent.  Following distillation at pH 9.5,  ammonia  concentration
can  be determined by U.S. EPA Methods 350.1, 350.2, or  350.3  or APHA Methods
417A,  417B, 4170, or 417G.   The methods should be  consulted to determine
which is most appropriate for available equipment, expected concentrations,
and  expected levels of precision  and accuracy.

     Calibration and Preventive Maintenance--Calibration procedures  should
follow those  specified  in the method.   If  samples  are  being  distilled,
standards should  also  be distilled  prior to analysis to check  for ammonia
contamination or loss during  processing.   Concentrations of the calibration
standards should bracket  the  sample  concentrations.  If a  sample concentration
is outside the range of  calibration,  then  an  additional calibration  standard
should  be analyzed  to  check  if the result  is  within the linear  range  of

                                      55

-------
                                                                  Effluent
                                                        Nitrogen  (Ammonia)

 the  method.  Alternatively,  tne sample  should  be diluted to within the
 calibration range  and then reanalyzed.

     Quality  Control  Checks—Duo! icate analyses  snould  be conducted on
 a minimum of 5 percent of the total  number  of samples, with  an  additional
 5 percent of the  samples spiked and analyzed  -or percent recovery.  A blank
 should be analyzed with  each batch of  samjles.   A U.S. EPA  performance
 sample should be analyzed at 1 :ast once per  quarter.

     Corrective Act Ion--Contamination  of ammonia  samples  .-an  occur easily
 due to the volatile  nature of anmorna.  To  prevent possible cross-contamination,
 reagents used for other analyses  that  contain ammonia (e.g.,  colorimetric
 phenol)  should be isolated  from samples  and standards used  for ammonia
 determinations.  In addition, cleaning  preparations that contain  significant
 quantities of ammonia (e.g.,  Pinesol,  wax removers) should  not be  used
 in the laboratory  area where ammonia determinations are performed.

     Contaminated glassware should be  rinsed with  1+1 HC1 and  then with
 distilled water.   To check  for  contamination,  blanks should  be  analyzed
 whenever a new reagent is prepared.

     Data Quality and  Reporting—Detection and accurate  quantification
 of ammonia in effluent is routinely  attainable,  although method detection
 limits can vary widely because of methods or instrumentation.  The  analytical
method should be consulted to determine  expected detection limits, precision,
 and  accuracy.  Data should  be reported in mg/L as N to a  maximum of three
 significant  figures.  Results should be  reported  for all determinations,
 including QA replicates and  spiked  samples.  Any factors that may have
 influenced sample  quality should also be  recorded.
                                     56

-------
                                                                  Effluent
                                                 Nitrogen (Total  Kjeldahl)

Nitrogen (Total Kjeldahl)

Field Procedures--

     Co11ection--Kjeldahl  nitrogen  samples can be  collected in glass  or
plastic containers.  Caps should be unlined,  as paper  liners and/or glue
may  interfere with the analysis.   Each  container and  cap should  be rinsed
thoroughly with sample water prior to  sample collection.

     Processing—If  possible, samples  should be  analyzed immediately after
collection.  If Immediate analysis is  not  possible, samples can be  stored
up  to  28 days by  acidification to pH<2  with  sulfuric  acid and refrigeration
at  40  c.   The length of delay before analysis should be recorded on the
log  sheet.
 ^
Laboratory Procedures--

     Analytical  Procedures—Approved  test  procedures for the analysis of
total Kjeldahl nitrogen  include U.S.  EPA  Methods 351.1, 351.2,  351.3,  or
351.4  and APHA Methods  417B, 4170, 417E,  420A,  or 420B.  The  methods should
be  consulted to determine which is most appropriate  for  available  equipment,
expected  concentrations, and expected levels  of  precision and  accuracy.

     Calibration and Preventive Maintenance--Calibration procedures should
follow  those specified In the method.  If samples  are being digested and
distilled, standards  should also be digested  and distilled prior to analysis
to  check for amnonia contamination or loss during distillation.   Concentrations
in   the calibration  standards should  bracket  the  sample --oncentrations.
If  a sample  concentration is  outside  of the  range of calibration, then
an  additional calibration  standard should be analyzed to check if  the result
is  within the linear  range  of  the method.  Alternatively, the  sample should
be  diluted to within  the calibration  range and  then reanalyzed.
                                     57

-------
                                                                 Effluent
                                                 Nitrogen  (Total Kjeldahl)

     Quality  Control  Checks--Duplicate  analyses should  be  conducted  on
a minimum of 5 percent of the total  number of  samples, with an  additional
5 percent of the  samples spiked and analyzed  for percent recovery.  A blank
should be analyzed with  each batch of samples.   A  U.S. EPA  performance
sample should be analyzed at least  once per quarter.

     Corrective Action—Because ammonia  is a component of Kjeldahl nitrogen,
precautions against contamination chat were described  for ammonia analyses
c ould be followed.

     Data  Quality and Reporting—The detection and accurate quantification
of Kjeldahl nitrogen  in effluent is routinely  attainable,  although  method
detection limits can vary  widely because of methods or instrumentation.
The analytical method  should be consulted  to determine  expected detection
limits,  precision, and  accuracy.  Data  should be reported in mg/L as N to
a maximum of three significant  figures.   Results should  be  reported for
all determinations,  including QA replicates, blanks, and spiked samples.
Any factors that may  have influenced sample quality should also be recorded.
                                     58

-------
                                                                 Effluent
                                            Nitrogen (Nitrate  and Nitrite)

Nitrogen (Nitra".e  and Nitrite)

Field Procedures—

     CpJJ_ec_tioin--Nitrate-nitrite  samples can be collected in  glass or plastic
containers.  Prior  to  sample  collection,  each  container  and cap  should
be rinsed thoroughly with sample  water.

     Processing--Nitr--, e-nurite  samples should  be  analyzed immediately
after collection.   If a  Jelay occurs,  samples can be  stored  for up to 24  h
by acidification to pH<2 with  sulfuric  acid  and refrigeration at  40 C.
Samples must not be  preserved using mercuric chloride because  the mercuric
ion accelerates the  degradation of the cadmium-reduction column (APHA  1985).

Laboratory Procedures—

     Analytical  Procedures—Approved test procedures for the analysis of
nitrate-nitrite include  U.S.  EPA  Methods 353.1,  353.2,  and  353.3 and APHA
Methods 418C and 418F.  The methods  should be consulted to determine which
is most  appropriate for  available equipment,  expected  concentrations, and
desired  levels of precision and accuracy.

     Calibration and Preventive Maintenance—Calibration procedures  should
follow  those specified  In the method.   Efficiency of each  reduction  column
should  be checked  by  comparing a nitrite standard  to  a  reduced nitrate
standard at the same concentration.  This  efficiency check  should be made
a I the  beginning  and  the end of each sample run  and at a minimum frequency
of every 10 samples.  Reactivate the copper-cadmium granules when reduction
 falls  below 75 percent.

      Concentrations of the calibration standards should  bracket the  sample
 concentrations. If a sample concentration is outside the range of calibration,
 then an  additional calibration  standard should  be  analyzed to check  if

                                      59

-------
                                                                  Effluent
                                             Nitrogen  (Nitrate and Nitrite)

 tne result  is  within the linear  range of the method.  Alternatively,  fhe
 sample  should be diluted to within  the calibration range and reanalyzed.

     Quality Control Checks--Duplicate analyses should be conducted  on
 a minimum of 5 percent of the total number of samples, with  an  additional
 5 percent of the  samples spiked  f. analyze for percent recovery.  A blank
 -.-at nas been run  through the reduction column  should  be  analyzed with
 each  batch of  samples.  A  U.S.  EPA performance sample should  be analyzed
 at  least once per quarter.

     Corrective Action—Various  components of the effluent can interfere
 with the analysis.  The method should be reviewed for ways to remove possible
 interferences prior to  analysis.   Possible intereferences include suspended
 solids, residual chlorine, oil and  grease, and  high concentrations of  iron,
 copper, or other metals.

     The  area where nitrate-nitrite analyses  are performed  should  be well
 isolated from exposure to nitric acid or nitric  acid fumes.

     Data Quality  and Reporting—The detection and accurate  quantification
 of  nitrate-nitrite  in  effluent are  routinely attainable,  although method
detection limits can vary because of methods or  instrumentation.  The analytical
method should be  consulted to determine expected  detection limits,  precision,
 and accuracy.  Data  should  be reported in  mg/L  as N to a maximum of three
 significant figures.   Results of all determinations should  be reported,
 including QA replicates, blanks, and  spiked  samples.  Any factors that
may have influenced sample quality  should also be  reported.
                                     60

-------
                                                                 Effluent
                                                       Phosphorus (Total)
Phcsphorus  (Total)

Field Procedures--

     Collection--Phosphorus  samples may  be collected  in glass or plastic
containers.   Containers should be  rinsed with  IN  HC1  followed by  several
r-.nses  with distilled water.   Detergents containing phosphate should never
be used on  containers or labware that  is to be used for  phosphate analysis.
Sample  containers and lids should  be ri sed thoroughly with sample water
before sample collection.

     Processing—Phosphorus  samples  can  be stored up to  28  days before
analysis by acidification to pH<2 with sulfuric acid and refrigeration
at 40 c.  Samples  with low concentrations of  phosphorus  should not be stored
in plastic  containers, as  phosphates may adsorb onto the container walls.

Laboratory  Procedures—

     Analytical  Procedures—Approved  test procedures for  the analysis of
total phosphorus  in effluent  are U.S.  EPA Methods  365.1, 365.2, 365.3 and
365.4 and APHA Methods 424C,  424F, and 424G.   The methods should be  consulted
to determine which is most appropriate  for  available equipment,  expected
concentrations, and desired levels of precision and  accuracy.

     Calibration and Preventive Maintenance—Cal ibration procedures  should
follow  those specified  1n the method.   Concentrations of the  calibration
standards should  bracket the sample concentration.   If a sample  concentration
is outside  the range  of calibration,  then  an  additional calibration  standard
should be  analyzed  to check  if the result is  within the linear  range of
the  method.   Alternatively,  the sample  should be diluted to  within the
calibration range and reanalyzed.

     Quality Control Checks—Duplicate analyses should  be conducted  on
a minimum of 5 percent of the total number of samples,  with an  additional
                                     61

-------
                                                                  Effluent
                                                        Phosphorus  (Total)

5  percent of the  samples spiked to analyze  lor percent recovery.   A blank
snould be  analyzed with each batch  of  samples.   A U.S. EPA  performance
sample should be analyzed  at least once per quarter.

     Corrective Action—Because  phosphorus contamination can occur from
a  variety of sources,  it is recommended that a clearly marked  set  of  lab-
v«are be  dedicated  to  only phosphorus analysis.  This laoware  should never
be exposed to phosphorus detergents or reagents containing phosphate.

     Various  components of the effluent can interfere wuh the analysis.
The method should be reviewed for ways to  remove  interferences or  adjust
for  interferences from  components that cannot  be removed.   Silica and arsenic
are possible positive  interferences,  while  hexavalent chromium  and  nitrite
can cause low recovery.

     For  highly colored or turbid  samples, additional  sample preparation
(e.g., further  oxidation  or filtration)  may be  required  prior to color
aevelopment.   In  any  case, blanks  should be prepared  by  adding all  the
reagents except  the coloring reagents  to the  sample.  Measure absorbance
in the  sample blank at the wavelength used  for the phosphorus determination
and subtract this  absorbance  value  from  the sample absorbance prior  to
calculation of phosphorus concentration.

     Data  Quality and Reporting—Detection and accurate  quantification
of total  phosphorus in  effluent is routinely attainable.  Actual  method
detection limits can vary because of  methods or instrumentation. The analytical
method should be  consulted to determine expected detection  limits, precision,
and  accuracy.  Data should be  reported  in mg/L as P to  a maximum of three
significant figures.   Results  of all determinations should be reported,
including QA replicates  and  spiked samples.  Any  factors that may have
influenced sample quality should  also be reported.
                                     62

-------
                                                                 Effluent
                                                 Priority  Pollutant Metals

P-iority Pollutant Metals

Field Procedures--

     Col lection--The  best containers  for collection of trace metal samples
are made of quartz or  7FE.   Because these containers are expensive,  the
preferred  sample container is made of polypropylene or linear  polyethylene
v»ith a polyethylene cep  (APHA 1985).   Boros il icate glass containers  may
be  used,  but do not use  soft  glass containers or container*  .nth aluminum
or cardboard lined lids.

     A minimum  sample size of 100  mL is  required for the analysis of all
priority pollutant metals,  except mercury, which requires an  additional
100 ml.  To allow for duplicates,  spikes, and potential reanalysis, a sample
size of 0.5-1 L  is recommended for the  entire  list of priority  pollutant
metals.
     Possible  problems  during sample  collection involve contamination from
the sampling device, airborne dust,  cross-contamination from previous samples,
and  loss  of metals by adsorption on container walls or precipitation after
collection.  Contamination can be  minimized by  avoiding the use  of metal
when  collecting effluent  samples.  Automatic sampling devices  should be
free of metal  parts in contact with the sample.  Prior to use,  containers
should be thoroughly cleaned with  detergent solution, rinsed  with tap water,
soaked in acid, and then rinsed with metal- free water.  Any glass or  plastic
parts associated with the sampling equipment should be cleaned  in the same
manner.   For quartz, TFE,  or glass containers,  use 1*1 HN03,  1+1  HC1, or
aqua  regia (3 parts cone HC1 + 1 part cone HN03)  for soaking.   For  plastic
material, use 1+1 HN03 or 1+1 HC1.  Reliable soaking conditions are 24 h
at  700 c  (APHA 1985).  Oo  not use chromic acid when preparing  containers.
For metal  parts, clean  as  stated  for glass  or plastic, except  omit the
acid  soak step of the cleaning  procedure.  All  acids should  be at  leas*
reagent grade.

                                     63

-------
                                                                  Effluent
                                                 Priority Pollutant Metals

     Processing—Acidify samples to pH<2 with  nitric  acid  immediately after
sampling.   Usually 1.5 ml concentrated HN03/L sample is sufficient.  Check
that adequate acid has  been  added by testing an  aliquot using pH paper
or a pH meter.  (Do not  insert any test materials  into the  sample container.)
Use high-purity  acid for preservation.

     Recommended maximum holding  t  ne  for trace  metal  samples  is 6 mo,
except for mercury samples, whicn should  be held a maximum  of 28 days  prior
to analysis (U.S. EPA  1982).

Laboratory Procedures--

     Analytical Procedures—Priority pollutant metals should be analyzed
according to the methods identified in Table IB (see Appendix A) of U.S. EPA
(1984).   Prior  to the  analysis, the effluent  sample must  be digested using
tne acids specified in the procedure.   The digestate can then be  analyzed
by  flame Atomic  Absorption Spectrophotometry (AAS), graphite furnace AAS,
or  Inductively Coupled Plasma  (ICP), depending  on the sample concentration
and required detection limit.  Mercury  analyses  must be performed on a
separate  sample  aliquot  by cold  vapor AAS.

      ICP can be  used to screen samples  for  elements  that  are present in
relatively  high  concentrations or for those that may  require more sensitive
analysis by  graphite  furnace  AAS.  Analysis by ICP can be subject  to inter-
element  interferences, while graphite furnace  AAS can be subject  to matrix
problems  from  acid  or salt content of  the  samples.  The detection  limit
of the selected method  should  be adequate to allow  determination of compliance
with water quality criteria.

      Calibration and Preventive  Maintenance—In  general, all  instruments
 must be calibrated daily and each time the  instrument is  set up.  Calibration
 procedures should follow those for the  specified  method  for each  analysis.
 Calibration standards must  be prepared  using the  same concentrations of
 acids as will result  in the samples following  sample preparation.
                                      64

-------
                                                                Eff uent
                                                Priority Pollutant Metals
     After an instrument has been  calibrated, the accuracy  of the initial
calibration should be verified by the  analysis of certified control  solutions
at a  frequency  of  once every 10 samples or  every 2 h during an analysis
run, whichever  is more frequent, and after  the last analytical  sample.
If a  certified  control  solution is not available, then a standard should
be used.   Analyte for this standard should be  from a different source  than
tnat  for  tne initial calibration.  If the deviation of the continuing cal i-
bratio verification  is greater than the calibration control limits  specified
in  the method,  then the instrument must be recalibrated, and the  preceding
10 samples reanalyzed.

     All  equipment  should  have  scheduled routine preventive maintenance,
and a record of all maintenance  should be  kept in  a  logbook.   Critical
spare parts should be kept on  hand.

     Quality Control Checks—To establish an estimate  of  precision, a minimum
of 5  percent of  the total number of samples  should be  analyzed in  duplicate
or  one  duplicate for each survey,  whichever is more  frequent.   When more
tnan  20 samples are  to  be  analyzed  for one  survey,  the project manager
may  choose to   implement a program  of  triplicate analyses.  The  overall
percentage of replicates should  be  at least  5  percent.  To establish an
estimate of recovery,  samples  spiked  before digestion should  be analyzed
at the same frequency  as replicates.  Spike concentrations to be added
should  be approximately equal  to  the  concentration found  in the unspiked
sample.   An acceptable  range  of  spike concentrations is 0.5-5  times  the
sample concentration.

      A method  blank should be carried through all  digestion and analysis
 steps at  a minimum frequency of every 20 samples,  or  for each batch of
 samples  analyzed, whichever is more frequent.   If the concentration of
 the blank is less than  the  required detection limit,  then no correction
of sample results  is  performed.  If the blank contamination is  extensive,
 then the  batch of samples associated with the  blank should be reanalyzed.
                                     65

-------
                                                                 Effluent
                                                 Priority Pollutant Metals

In general,  the blank should  be  10  times less than  the concentration of
the analyte in the sample.   The  data should be corrected  by data users
for  the  blank values between  the  required detection  limit and the control
1imit.

     For  ICP  analysis,  additional  QC checks should  include an interference
cneck sample to ;enfy  interelement and background  correction  factors.
For  graphite furnace AAS, additional  QC checks should  include duplicate
injections with the mean value repo. ed.  Relative  standard deviation of
the  readings should be within control  limits, or the sample must be rerun
at least once.

     Corrective Action—If the concentration of the  field  or method blank
is greater than the required detection limit, then all  steps in the sample
handling should be  reviewed.   Many  trace metal contamination problems are
due  to airborne dust.  Keeping containers closed and rinsing all  handling
equipment immediately prior to use minimizes dust problems.

      Poor duplication may be  caused by inadequate  mixing of the sample
before removal of aliquot,  inconsistent contamination,  inconsistent  digestion
procedures, or instrumentation problems.

      Poor spike recovery may  be  caused for  the same reasons  as poor du-
plication.  However,  if  duplicate results are acceptable, then  poor spike
recovery may be  caused by loss of  analyte during  digestion or by sample
matrix interferences during analysis.  To check  analyte loss during digestion
or  for  low  recovery due  to  interferences during  analysis, spike the sample
after digestion and compare  the  analysis to the predigestion spike.   If
 the results  are different,  then the digestion technique should be adjusted.
 If the results  are the same  then  dilute the sample  by  at  least  a factor
of  five  and  reanalyze.  If spike recovery is still poor,  then standard
 additions, matrix modifiers,  or another method is  required.
                                      66

-------
                                                                  Effluent
                                                  Priority Pollutant Metals

     Data Quality and Reporting—Data should  be reported to a maximum of
three significant  figures as mg/L.  These  units are  consistent with  ODES
format.   Detection limits can vary widely because of methods or instrumen-
tation.   The analytical method should  be  consulted  to  determine  expected
detection limits, precision, and accuracy.   The data report should include
duplicate, spike, and blank results.   The data should not be blank-corrected.
It  is  recommended that all new data included in  the ODES database  be blank-
corrected by the data  users.  The data nummary from the laboratory  should
also include:

     •    Digestion procedures

     t    Instrument detection limits

     •    Detection limit  for each element

     •    Blank associated with sample

     •    Deviations from  the prescribed method

     t    Problems associated with analysis.

For  a more  thorough QA review, additional  documentation  (e.g.,  calibration
curves)  may be requested.
                                      67

-------
                                                                  Effluent
                                       Priority Pollutant  Organic Compounds

Priority Pollutant Organic  Compounds

Field Procedures--

     Col lection--The priority pollutant organic compounds can be separated
into purgeable (volatile) and extractable (base/neutrals, acids, pesticides,
ani PCBs) compounds.   Container preparation and collection techniques differ
fcr these two Croups.

     If a volatile analysis is required, two separate  40-mL glass containers
should be filled leaving no headspace.   The container, screw cap, and septum
should  be washed  with detergent, rinsed  once with  tap water, rinsed once
with  distilled water, and dried at >105o c.  Use of a solvent rinse will
interfere with the analysis.  To obtain  a sample  with no  headspace,  fill
the  vial  to overflowing  so  that a convex  meniscus forms at the top.  With
the cap liner's TFE side down, place the  cap  carefully  on  the opening  of
the  vial, displacing the  excess water.   Once sealed, invert the bottle
to verify the seal  by  demonstrating the  absence of air  bubbles.  Samples
for  volatile analyses should  be grabs  only, because many of the volatile
compounds of interest may be lost while  compositing.

     For  the extractable analyses, amber glass  containers with TFE-lined
screw caps should  be used.  Clear glass  may  be used if the  sample is  kept
from  light  (i.e., stored  in  an ice chest  or refrigerator).  It is advised
that 4 L be collected  to allow for reanalysis  if necessary.   Eight L  should
be collected for samples to be analyzed  as spikes or duplicates.  The container,
lid, and liner should  be washed with detergent, rinsed twice with tap water,
rinsed  once with  distilled  water, and the liner  and container rinsed once
with high-purity methylene chloride or  acetone.  Kiln drying of the container
at  4500 c for at  least  1  h  can be substituted  for solvent rinse.   There
is evidence  that  some priority pollutants  (e.g., hexachlorobenzene)  in
aqueous solution adsorb strongly to TFE  lid  liners,  ."his may be a potential
problem if the effluent contains ".hese  pollutants.

                                     68

-------
                                                                 Effluent
                                      Priority Pollutant  Organic Compounds

     Automatic  sampling  equipment  should  be clean, and  free of plastic
ana other sources of contamination.   Ideally,  a  sample bottle  containing
"organic-free water" carried  through all processing and  handling will serve
as a field blank.  Because organic-free water  is not readily available
outside of a  laboratory, this  may be  impractical for all  monitoring programs.
Alternatively, an  empty sample jar carried through all processing  and handling
will serve as a  field blank.   A solvent rinse of the bottle  should be analyzed.
The solvent should be the same as  that used  for  sample extraction.   This
will  serve as a check on contamination tha' may occur  curing  shipping and
storage.

     Process ing--Samples should be stored in the dark at 40  c or on  ice.
Freezing of the  samples in transit should be  avoided as the water expansion
causes  broken vials.  For volatile organic compounds, maximum recommended
holding  time is 7  days.  For extractable organic compounds, maximum recommended
holding time is 7 days  prior to  extraction, with  all analyses  completed
within 40 days of extraction.

     Sample extracts should be stored frozen  at <-10o c.

Laboratory Procedures--

     Analytical  Procedures--Priority  pollutant  organic compounds  should
be  analyzed according to the methods  identified  in Tables  1C  and 10  (see
Appendix  B) of U.S. EPA  (1984).   If samples with analyte levels that differ
by  several orders of magnitude  as determined by  screening or experience
are being analyzed together,  carryover  can be a problem.  This should  be
kept  in mind when reviewing data obtained by auto-injection.  For volatile
organic compounds, samples  that appear  "clean"  should be analyzed first.
Analysis of  an appropriate solvent after a high-level  sample can minimize
the chance  of  false  positives.   If separate injection syringes are used
 for standards and samples, any bias  between syringes  should be accounted
 for.  They  should  be  cleaned  thoroughly  between each use to avoid cross-
contamination.
                                     69

-------
                                                                 Effluent
                                      Priority Pollutant Organic Compounds
     Detection  limits vary with specific  compounds and methods  but  are
usually obtainable  in  the  range of  5-10 ppb  for base,  neutral, and acid
compounds (U.S. EPA  Method 625), 0.005-0.10 ppb for pesticide/PCB analysis
(U.S. EPA Method 608), and 1-10 ppb  for volatiles (U.S. EPA Method 624).
Sample  size, final volume,  co-extractives and nature  of effluent can  all
affect the actual detection limit.   Detection  limits , equested by  project
managers should take  into  account obtainable  levels  as well as required
301(hj program criteria.

     Calibration and  Preventive Maintenance—Before  beginning analysis
of samples, a calibration curve bracketing the working range must be  performed.
This calibration  should be repeated  after all major equipment disruptions.
Calibration checks of the  6C/MS system should  be done at the  beginning
and  end of each day and at least every 12 h to verify that the instrument's
response is in control.  Specific instrument tuning criteria (e.g.,  DFTPP,
BFB) are provided in each method (U.S. EPA 1984).  Calibration of  the GC/ECO
system should be done  at the beginning of each  day and verified at  least
every 6 h.  These  are only  recommended minimum frequencies.   Depending
on the nature of the samples, it may be necessary to verify calibration
more frequently.

     A routine QC check on each lot  of analytical  reagent used  in  extraction
can  prevent undetected contamination problems.  Each lot of alumina, silica
gel  , sodium  sulfate, or  Florasil  used should  be monitored  as  a  possible
source  of  contamination.  The efficacy of adsorbants often  varies  between
lots and  should be routinely monitored.   Each lot of surrogate mixture
should  also be checked for contaminants.

      Equipment  should be maintained and  serviced  routinely by  experienced
chemists according to manufacturers instructions  and good laboratory  practices.
Logbook records should  document maintenance  for  each measurement device.
                                     70

-------
                                                                 Effluent
                                      Priority  Pollutant Organic Compounds

     Quality Control Checks--To  monitor reproducitnlity and  extraction
efficiency, duplicates and matrix spike  analyses should be  performed.
A frequency of 5 percent each or one each per  survey (whichever is greater)
is the recornnended minimum.  When more than  20 samples are  to be analyzed
for one  survey,  the project manager may  choose  to implement a program of
triplicate  analyses.  The overall  percentage of  replicates  should be  at
least 5 percent.  Method interferences can be caused by  contaminated glassware,
reagents, solvents, or processing  hardware.  To check  for ;ontamination,
one method blank should be processed  after each group  of  Z  samples or
with each batch, whichever is  greater.  Addition of known amounts of surrogate
compounds  to  each  sample before purging or extraction  will serve to monitor
preparation and analysis  of  samples.   A spike  prepared in the  field  can
be submitted to the  laboratory as another measure of accuracy.

     Corrective Action—When  results of QC samples fall outside of established
limits, several courses  of  action are available.   Contamination  in  the
lab reagent blank sample is  cause  for positive findings of  the  same compounds
in samples to  be  suspect.   If contamination  is  extensive, reanalysis of
the  entire associated group may  be in order.  Blank contamination should
be kept to less than 10 percent  of  sample values  and  preferably below  the
method detection  limits.  Contamination found  in  the field  blank should
be considered  when  looking at the  associated sample data.  Extensive contam-
ination of lab or  field blank  (>30 percent of sample  values)  should  lead
to  a  detailed  review of  laboratory, sampling,  transport, and  storage pro-
cedures.  Phthalates, methylene chloride,  and toluene are common laboratory
contaminants  that  may  be detected in  blanks above the method detection
 limit.

      Poor duplication may be caused  by inadequate  mixing of  the  sample
 before reroving aliquots, inconsistent contamination, inconsistent extraction
 procedures,  or  instrument problems.  If further  replication of an analysis
 produces poor results, a step-by-step examination  of the  method may be
 necessary to determine the reason for the poor results.  If the results

                                     71

-------
                                                                  Effluent
                                       Priority Pollutant Organic Compounds

of the duplicates fall  outside control limits, the  laboratory should contact
the data user and discuss the possibility of further  analyses.

     Poor spike recovery may be caused for the same reasons as poor duplication
or by matrix effects.   If the spiked compound is added  at  a concentration
much less than that found in the sample, recovery may be difficult to deter-
mine.  This problem is  difficult to  avoid, as most  environmental  samples
cor:ain  unknown  concentrations of organic compounds.   To check for analyte
loss curing processing,  a  step-by-step examinatior  of  c.ie method using
a spiked blank is necessary, with measurements of the .nalyte at each step.

     Sample results that  fall  outside  the  established calibration curve
are suspect until linearity of response can be shown  at  that concentration,
or  the extract diluted  appropriately and  reanalyzed.  Extremely high concen-
trations of organic compounds may  saturate the  extraction capabilities
of  the method  and may necessitate  re-extraction of  a smaller sample size
or  use of a more appropriate method.

      If the instrument's  continuing calibration (single point) falls outside
control limits, no samples should  be analyzed until  the calibration  is
within control  limits.   The standard should be  reinjected to confirm the
problem and to discount the possibility of operator error.   If still outside
of  control limits, the instrument should  be  recalibrated (multi-point)
and at least the  previous  sample reanalyzed and  results compared.   This
may indicate that  reanalysis of  all  samples since last  calibration is unneces-
sary.

      Data Quality and Reporting—A  data summary  for each sample should
be  submitted.   All  data should be reported  in  ug/L using two  significant
figures.   Data should not be  blank-corrected.  Spike recoveries,  relative
percent difference between duplicates, and  blank results (ng/sample) should
also be  submitted.  The following information is also needed for each  sample
to  allow  independent QA review:

                                      72

-------
                                                                   Effluent
                                       Priority  Pollutant Organic Compounds


     •    Sample  volume extracted


     •    Final volume of extract


     •    Amount  of  extract  injected


     •    Instrument detection  limits


     •    Detection  limit  f:r  each  compound


     •    Blank associated with sample


     •    Deviations from  the  prescribed  method


     t    Problems associated  with  analysis.


For a more thorough QA review,  additional  documentation (e.g., chromatograms,
computer listings) may be  requested.
                                       73

-------
                                                                 Effluent
                                         Total and Fecal  Coliform Bacteria
Total and Fecal  Colifortn Bacteria

Field Procedures--

     Col 1ection--Samples should be collected  in clean, sterile polypropylene
or glass containers.  The sample containers must ce resistant to sterilizing
conditions an-- to the solvent action of water.   The  container lids must
not  produce bacteriostatic  or nutritive compounds upon sterilization.
The  sample containers must  sea1 tightly.  Containers  with chips, cracks,
or etched marks should be discardeo.
     Heat-resistant glass or plastic  sample containers should be autoclaved
at 1210 c for 15  min.  Alternatively,  dry glass containers  can be sterilized
in a  hot-air oven at 1700 c for at least  2 h.  For plastic  containers that
are not heat-resistant,  ethylene oxide gas  sterilization is acceptable
(Bordner et al.  1978).  Containers sterilized by gas should be stored at
least 12 h before use to ensure all  gas  has dissipated.

     If  the  sample water has residual chlorine, sodium thiosulfate should
be added  to neutralize the chlorine and  thereby  prevent continued bactericidal
action  after sample collection.   In this manner, the true  microbial content
of the water at  the time of sampling  can be estimated more accurately.
If sodium thiosulfate must be  added  to  a  sample, it should be added to
the sample container prior to sterilization so that the final concentration
in the  sample will be 100  mg/L.  For a  120-mL container,  0.1 ml of a 10-
percent  solution  of sodium thiosulfate will neutralize a sample containing
as much  as 15 mg/L  of residual chlorine  (APHA  1985).

      If  the  sample water contains heavy metals in concentrations exceeding
0.01 mg/L, a chelating  agent should  be added to the sample container to
reduce  metal  toxicity.   This is particularly important if  samples are not
analyzed within 4 h after collection.  APHA (1985) recommends  using the
disodium salt of  ethylenediaminetetracetic acid (EOTA), adjusted to pH

                                     74

-------
                                                                 Effluent
                                         Total  and  Fecal Coliform Bacteria

6.5, and  added to the sample contaiier before sterilization.  For a 120-mL
container, addition of 0.3 ml of a 15-percent EDTA solution is considered
adequate (APHA  1985).

     It is critical that samples are not contaminated  during the collection
process.   To  avoid contamination,  sterilized  containers  should  be  kept
sealed  until  they are used, containers should be  filled  without rinsing,
ana container  lids should be  replaced  immediately after the samples  have
been collected.  When remo- id from  containers, lids  should be held face
down in one hand and not set  «iown on any surface.  Adequate headspace (at
least  2.5 cm)  should be left  in each sample container to facilitate mixing
prior to analysis.

     Processing—Samples  should be analyzed as soon as possible after col-
lection.   If  a delay occurs, samples should be held at 40 C  for a maximum
of 6 h.  The length of delay  should be noted on the log sheet.

Laboratory Procedures--

     Analytlcal Methods—Details of the membrane  filter (MF) method and
tne most  probable number (MPN)  method  are  presented in Part  III  of  Bordner
et  al. (1978)  and in  Parts 908  and  909 of  APHA  (1985).   Although the MF
method is more precise  than the MPN method,  it is also more  sensitive to
interference  from turbidity  in  samples.   Because the MF  technique usually
yields low and variable recovery from  chlorinated waters, the MPN technique
should be used for chlorinated effluent.

     Calibration and  Preventive  Maintenance—This  information is  reviewed
extensively in Part IV of Bordner et al. (1978) and  in Part 902 of APHA
(1985).

     Quality  Control  Checks — Qua! ity control checks  for  col i form analyses
are listed  in  detail  in Part IV of Bordner et  al.  (1978)  and in  Part 902
of APHA  (1985).  The list  includes:
                                      75

-------
                                                                  Effluent
                                          Total  and Fecal  Coliform Bacteria
     •    Sterility  checks on media, dilution and rinse water, glassware,
          and membrane  filters

     •    Duplicate  analyses on 10 percent of samples  and  on at least
          one sample  per  test run

     •    Colony verifications on a monthly basis.

     Corrective Action—Procedures detailed  in the relevant sections o^
Bordner et al. (1978) and APHA (1985) should be followed.

     Data Quality and Reporting—Table 909:11 of APHA  (1985) presents 95-
percent confidence limits for the  MF  method for coliform colonies  of 1,
2, 3, 4, 5 and 10.  The precision of the MPN method increases with increasing
nuntoer of replicates.   With  five tubes, each with 1 mL of  sample, a completely
negative result  is  expected less  than 1 percent of the time (APHA 1985).
Confidence limits (95 percent) for various MPN  counts are given  in  Tables
908:111, 908:IV, and 908:V of APHA  (1985).

     Using  the MF method, data  should be reported as densities of col i forms
per 100 ml.   Using the  MPN method,  data should be reported as MPN  values
per  100 ml.   Results  should be reported for  all determinations, including
QA replicates, sterility checks,  and  colony  verifications.  Any  factors
that may have  influenced  sample  quality  should  also be reported.
                                      76

-------
                                                                 Effluent
                                                     Enterococcus Bacteria
Eiterococcus Bacteria

Field Procedures--

     Collection—Samples should  be collected in clean, sterile polypropylene
or glass containers.  The sample containers must be resistant to sterilizing
conditions  and to the solvent  action of water.   The  container lids must
not  produce bacteriostatic or  nutritive compounds upon sterilization.
Tne  sample  containers must seal  tightly.  Cntainers  with chips, cracks,
or etched marks should be discarded.

     Heat-resistant glass or plastic sample containers should be autoclaved
at 1210 C for 15 min.  Alternatively, dry glass containers  can be sterilized
in a hot-air oven at 1700 C for at least 2 h.  For plastic containers that
are not heat-resistant, ethylene  oxide gas  sterilization is acceptable
(Bordner  et al.  1978).  Containers  sterilized by gas  should be stored at
least 12 h before use to ensure  all gas has dissipated.

     If the  sample water  has  residual chlorine, sodium thiosulfate should
be added to neutralize the chlorine and thereby prevent continued bactericidal
action  after sample collection.  In this manner, the true  microbial content
of the water at the time of sampling can be  estimated more accurately.
If sodium thiosulfate must be  added  to a sample,  it  should be added to
the sample container prior to sterilization so that the final concentration
in the  sample Mill  be 100 mg/L.  For a 120-mL container, 0.1 ml of a 10-
percent solution of sodium thiosulfate will neutralize a sample containing
as much as 15 mg/L of residual chlorine (APHA 1985).

     If the   sample water  contains heavy metals in concentrations exceeding
0.01 mg/L, a chelating agent should  be added  to the sample container to
reduce  metal toxicity.  This  is particularly important if samples are not
analyzed within 4 h after  collection.  APHA  (1985) recommends using the
disodium salt of ethylenediaminetetracetic  acid (EOTA), adjusted  to pH

                                    77

-------
                                                                 Effluent
                                                     Enterococcus Bacteria

6.5,  and  added to the sample container  before sterilization.   For a 120-ml
container, addition of 0.3 ml of a 15-percent  EDTA solution  is  considered
adequate (APHA 1985).

     It is critical that samples are not contaminated during  the collection
process.  To avoid contamination,  sterilized  containers should  be  kept
sealed  until  they are used,  containers  should oe filled without rinsing,
ana container  lids snould be replaced  immediately after  the  samples  have
uten  collected.  When removed  from containers,  lids should be heU face
down in one hand and not set down on any  surface.   Adequate  headspact (at
least  2.5 cm)  should be left in each sample  container to facilitate mixing
prior to analysis.

     Process ing--Samples should be analyzed  as soon as possible after col-
lection.   If  a delay occurs,  samples should be held at 40 C for a maximum
of 6 h.  The length of delay should be noted on the log sheet.

Laboratory Procedures--

     Analytical Methods—Methods for analyzing enterococcus bacteria are
currently being finalized by U.S. EPA.

     Calibration and Preventive Maintenance—This information is  reviewed
extensively in Part IV  of Bordner  et al.  (1978)  and  in  Part 902  of APHA
(1985).

     Quality  Control Checks—Qua! ity  control checks for these analyses
are  listed in detail  in Part IV  of  Bordner et al. (1978)  and in  Part 902
of APHA (1985).  The  list includes:

     •     Sterility checks on media, dilution  and  rinse water, glassware,
           and membrane  filters
                                     78

-------
                                                                  Effluent
                                                      Enterococcus Bacteria
     t    Duplicate analyses  on 10 percent of samples  and  on at least
          on-»  sample per test run

     0    Colony  verifications on a monthly basis.

     Corrective  Action — Procedures detailed in  the  relevant  sections of
3c"dner et al.  (1978) and APHA (1985) should be followed.

     Data Quality and  Reporting—Data should be reported  according  to the
specifications currently being finalized by U.S. EPA.
                                      79

-------
                   MONITORING  THE RECEIVING ENVIRONMENT
     Recommended  methods for measuring receiving-environment  variables
during  301(h) monitoring programs are described in this  section.  The initial
major  section includes general procedures for sampling  preparation, sampling,
and sample handling.  The section on sample  handling contains tables  that
sjmmarize the recommended coTaction specifications and analytical methods
for specific receiving-environment  variables.   The second major  section
presents detailed  procedures  for measuring 21 recp n'ng ^.•.vironment variables.

GENERAL METHODS

Sampling Preparation

     Prior to each field survey,  a sampling and analysis  plan  should be
prepared.   It should summarize all of the elements essential  for conducting
the  survey.  The  chief  scientist  or  a designee should  thoroughly review
the plan  (including QA/QC criteria)  before  each cruise, and ensure  that
its  essential elements  are understood by  all  members  of the  scientific
party.  The review should focus  on  the completeness  of the plan and the
clarity of  its objectives.  A complete sampling and analysis plan should
contain the  following major  elements:

     •    Identification of scientific  party and the  responsibilities
           of  each member

     •    Statement and prioritization  of study  objectives

     •    Description of survey area,  including  background information
           and station locations

     •    Identification of variables  to be measured  and corresponding
           required  containers and preservatives

                                      80

-------
     t     Identification  of all sample  splits  or  performance samples
          to be submitted with the samples

     •     Brief  description of sampling methods, including positioning
          technique, sampling  devices, replication, and any  special
          considerations  (including  handling  and  shipping hazardous
          materials)

     •     Detailed  cruise schedule, including time, date, and  location
          of embarkation and debarkation

     0     Identification of  onshore  laboratories to which  samples
          should be shipped after cruise completion

     •     Survey vessel  requirements (e.g., size,  laboratory needs,
          sample storage needs)

     •     Location and availability of an alternate  survey vessel

     t     All  special equipment needed  for the survey  (e.g., camera,
          nets,  communication  devices)

     •     All  pertinent QA/QC  procedures.

     It is essential  that  the  study  objectives and their  prioritization
be understood  by all members  of the  scientific party.   This will  ensure
that if modifications of the  survey plan become necessary  in  the  field,
their impact on  the overall  goals of the  cruise  can be evaluated  adequately.
After  the sampling plan  has been  reviewed, contingency plans should  be
outlined.  These  plans should  include potential problems and  their  solutions.
Possible  solutions to some problems  frequently encountered  during offshore
surveys are listed  in  Table  4.   Development  of contingency  plans can  be
greatly assisted by reviewing cruise  summary reports and  consulting  with
chief scientists of previous cruises.
                                     81

-------
       TABLE  4.   EXAMPLES  OF PROBLEMS  FREQUENTLY ENCOUNTERED  DURING
         OFFSHORE  SURVEYS AND RECOMMENDED SOLUTIONS  TO EACH PROBLEM
                  Probh
          Recommended Solution!s)
Sampling  equipment fails or is lost overboard
Sampling  efficiency reduced by poor * j'.
Sampling  efficiency 'educed aecause of seasickness
Sampling  delayed because  vessel is inoperative
A sample is partially lost or slightly contam-
inated

Bottom sampler  will not penetrate to a sufficient
depth after repeated casts
One  or more water bottles  fall  to trip  on
a particular cast
- Have necessary tools onboard to make  repairs
- Maintain  spare  parts  inventory for major
  equipment
- Have back-up equipment  onboard
- Have SCUBA equipment and divers onboard
  for retrieval
- Know where nearest tools, back-up equipment.
  or divers  are located on shore

- Extend cruise length
- Reschedule cruise

- Ensure scientific crew  is large enough
  to compensate for reduced personnel
- See solutions to second problem

- Reschedule cruise
- Extend cruise length once vessel is repaired
- Charter an alternate vessel

- Discard and  take another sample
  Add weight to  the sampler
  Move  a  short distance  before taking next
  sample

  Discard  all  samples  from  that cast and
  take a new cast
                                            82

-------
     The captain  of  the  survey vessel  should be  provided  with a copy of
*.?ie survey  plan to ensure  that it is consistent with the  equipment and
capabilities  of the vessel.   Modifications to the ship or cruise plan may
se required.

     To ensure that all required sampling equipment and supplies are available
at the time of sampling,  an  equipment checklist should  be constructed.
Spare  parts  and backup supplies  should be  included  in the inventory.  All
equipment should be  inspected before the  cruise, with sufficient time  allowed
•o make  necessary repairs or  replacements.   At the end of  each  cruise,
eauipment should be inspected, cleaned,  and stored properly.

Sampling Procedures

Station Location-

     Accurate navigation is essential  to ensuring  that stations  can  be
plotted and reoccupied with a high  degree of  certainty.  Because compliance
with 301 (h) regulations usually requires that  stations oe located at  specified
distances relative to a  fixed-point discharge  (e.g., within-ZID, ZID-boundary)
or  that stations be  sampled seasonally or  annually, failure to accurately
locate or relocate stations can strongly affect  the  results and conclusions
of a survey.   For  example, if a within-ZID  station is  positioned erroneously
outside the ZID, the resulting measurements may suggest  that  the environmental
impacts of the discharge are much less than  they actually are.  Alternatively,
if a station thought  to  be outside the  ZID is located mistakenly  within
the  ZID, conditions  would  appear to  be  worse  than they really are.  As
an additional example, failure  to accurately relocate  a  station may produce
results that  suggest conditions are  improving, when  they  actually may be
deteriorating.

     Although a variety  of  navigation and/or position  fixing systems are
available  currently,  factors  such as  price,  availability, and accuracy
vary considerably among  them (Tetra Tech 1986c).   The station positioning
system selected for a given survey  should  be able  to -neet  all regulatory
requirements  for  accuracy and should, at a minimum, provide  a high degree

                                     83

-------
of  precision  (i.e.,  repeatable measurements).   Positioning systems  that
are  precise but lack  a high degree of  accuracy may be  used after actual
station locations are  determined by  accurate,  independent means  (i.e.,
"ground-truthed").   For  bottom-related samples,  all positioning systems
snould be used  in conjunction with a  fathometer to ensure that sampling
occurs at the proper water depth (allowing for  tidal stage and any  fathometer
corrections).

Water  Sampling--

      For most  301(h) monitoring  programs, *ater column variables will be
sampled usi'i  water  bottles, in sit-j  instrumentation, or  a combination
of both of t ese techniques.

     Water  Bottles—Water bottle  samplers are  relatively simple devices
that generally consist of some type of  cylindrical tube with stoppers at
each end and a closing device that is activated  by a messenger  or an electrical
signal.  The most ccnmonly used samplers  of this description are the Kemmerer,
Van  Oorn,  Niskin, and  Nansen  samplers.   Each device samples  a discrete
parcel  of water at any  designated depth.   Frequently, multiple water samplers
are  fixed  on  a  rosette  frame  so that several  depths can be sampled during
one cast or replicate  samples can be taken at  the  same depth.

     Prior  to  deployment, the  stoppers of water  bottle samplers  are cocked
open on the sampling  vessel.  At this step, it is  critical that the interior
of  the sampler  and  stoppers remain free from contamination. All members
of the sampling team  should therefore  avoid  touching the  insides of the
sampler and stoppers,  and all samplers  and stoppers must be rinsed  thoroughly.

     After  cocking,  the sampler  is  lowered to  a Designated depth.  The
sampler must be open  at both ends r>o that water  is not trapped within the
device as  it  is being  lowered through the water column.  Once the sampler
reaches the desired depth,  it should be allowed  to equilibrate with ambient
conditions for 2-3 min before being closed.
                                     84

-------
     After equilibration,  the  closing device can be activated  by messenger
or electrical  signal, and the  sampler can be retrieved.  It  is  recommended
tnat  at  least two  samplers  be used  simultaneously for each  depth.   A  second
sampler will  provide a backup to  the primary  sampler in case  the  latter
device misfires or won't  trigger.  This will  eliminate  the  need  for  an
additional  cast.  A second sampler will also supplement the  primary  sampler
if the volume collected by the latter device is too small  for  all  required
subsampl ing and rinsing.  To  ensure that  all  subsamples at a  particular
depth are  collected from  the  same water  parcel, it is essential  that  they
all  be taken from a  single cast.  Multiple casts using a single water sampler
will  not  meet this objective.   Sample water must therefore be  used conser-
vatively after collection.

      Once  the water  sampler is brought  on board the sampling vessel, the
stoppers should be  checked  immediately for complete  seals.   If a stopper
is not properly sealed, water from the  sampled depth may have leaked out
upon  retreival and  been  replaced by  water  from shallower depths.   Because
this  kind  of contamination can  bias results,  the entire water sample should
be rejected.

      Accepted water samples should be subsampled  as soon as possible because
appreciable delay may result  in  unrepresentative  subsamples.   For  example,
measurement of variables  sensitive to biological  alteration (e.g.,  dissolved
oxygen, turbidity,  color, nutrients) or  settlement within the water sampler
(e.g., total  suspended solids,  settleable  solids,  phytoplankton)  can be
biased substantially by  subsampling  delays.

      In  Situ Instrumentation—A wide  variety  of  instruments capable of
measuring  water column variables in situ  are available.  Most are  deployed
from  the sampling vessel  using  a cable.   Sensors  housed within the instruments
measure the variables of  interest and transmit  data  in the form of electrical
signals back  to recorders on  the survey vessel.   The simplest instruments
measure conductivity  (i.e., for conversion to salinity),  temperature, and
water pressure  (i.e.,  for conversion  to depth).   Additional sensors and
instrumentation can be  included to measure a variety of  additional  water-
column variables  such  as  dissolved oxygen, pH,  transmissivity  (i.e., an

                                     85

-------
index of turbidity), oxidation-reduction potential, and specific ions  (e.g.,
a/rnonia).   Generally, the operating manuals  supplied with these instruments
provide detailed descriptions of how  to  calibrate, operate, and  maintain
the equipment.  If a particular manual lacks sufficient detail, the manufacturer
snould be contacted  for  specific guidance.

     Although the instrument  operating manuals  should be consulted  for
specific instructions,  several general  procedures for  operating j_n situ
instruments apply to  all  or most devices,  and hava a direct influence on
data quality.  When  acquiring  in  situ  instruments for  use  in marine  and
estuarine waters the following  features are  highly  recommended:

     •    Instruments  should  be  of  rugged constru"tion,  corrosion-
          resistant, and waterproof

     0    Instruments  should  be  capable  of operating with acceptable
          accuracy within the range of expected environmental  condi-
          tions

     t    Cables should  be  of adequate length and  strength

     •    Electrical connectors should be  easy  to use, waterproof
          when connected, and capable of being locked after connection

     •    External  sensors should  be protected  by housings  or other
          means

     •    Instruments  should  be easy  to calibrate on  board  the survey
          vessel

      •    Ideally,  a  service  center should be located  nearby so  mat
          an  instrument can be repaired  rapidly, if necessary

      •    If  sample contamination may be a  problem, all  sampling equipment
          should  be made of non-contaminating material.
                                      86

-------
     When using _l£ situ  instruments it  is  critical that  it be protected
from rough handling and adverse  environmental conditions.  The  following
precautions  should always  be  taken:

     •    Instruments  should be transported  in specially designed
          shipping boxes

     •    Instruments should oe  surrounded by a "birdcage"  when being
          deployed.   Frequently, instruments  are attached within a
          rosette frame when they  are used in  conjunction  with water
          bottles.  Caution should be  taken  to ensure that  the  "birdcage"
          or rosette frame do not  create sampling artifacts

     •    Instruments should be  securely lashed  down  in a safe area
          when on deck

     0    Instruments should be  rinsed with fresh or distilled water
          after each submersion

     •    Optical surfaces  should  be cleaned  with  alcohol and lens
          tissue  after each submersion

     •    Instruments should be  protected  from direct  sunlight and
          excessive heat, as plastic components  may be damaged  by
          heat

     •    External sensors  and  optical  ports  should be covered  and
          protected whenever  the instrument is not being used.

     When operating  in  situ instruments, the  following procedures  should
be followed to ensure that instruments  are prepared, deployed, ana retrieved
properly:

     •    Instruments should  be allowed  to warm up for a sufficient
          length  of time prior to calibration or  deployment
                                     87

-------
     •    Instruments should  oe field-calibrated at the  beginning
          of each day  of sampling.  All  circuits should  be tested
          at  the  same time.   Calibration  should be conducted  more
          frequently  if equipment malfunctioning  is suspected

     •    Upon  instrument deployment, the  survey vessel  should be
          anchored or drifting slowly

     •    Instruments should be deployed relative to vessel construction
          and sea  conditions  so  that cables  will not tangle  in the
          propel lor or  rudder assemblies

     •    When measuring  continuou: profiles, the lowering speed through
          the water column should not exceed  the equilibration  rate
          of the sensor having the slowest  response time

     •    Excessive  strain  should  not  be placed  on  the cable(s), as
          it could disrupt electrical connections.

     Routine maintenance  and inspection of  in situ instruments should follow
the manufacturer1 s recommendations.  General  procedures include:

     •    All rubber parts  of underwater connectors  should  be coated
          with silicone grease to ensure proper lubrication

     •    Plugs should  be inspected  for  bent  or  broken  pins, which
          may cause faulty connections and  flooded cables

     •    Cables should  be inspected  for nicks,  cuts, abrasions, or
          other signs of  physical damage

     •    Seals should  be inspected and periodically cleaned  and greased
          to ensure a waterproof  fit

     •    Oesiccant  should be inspected  and replaced with fresh or
          reactivated desiccant when necessary.

-------
     Factory inspection and  recal ibration  at  reccrrniended intervals is essential
to ensure that an in situ instrument  is  functioning  properly  and  will  continue
to  function properly during  future  cruises.   It  is recommended that  factory
service be conducted at least once  per  year.   Factory service  should always
be  conducted when  instrument malfunctions cannot be corrected by  following
the operating manual.  Factory service may  also  be  required  when part  of
an  in  situ system  is replacec, as all  components  are not  interchangeable
without factory recalibration.

     Log  Sheet — Variables  that should  be  recorced  on the  water sampling
log sheet are:

     •    Geographic location
     0    Date and time
     •    Weather  conditions
               Sea state
               Sky state
               Precipitation
     •    Station  number
     •    Sample number
     t    Replicate number
     t    Position coordinates
     •    Total depth
     •    Sampled  depth(s)
     •    Sampler  description
               Kind
               Volume
               Rosette, paired,  single
     •    Kinds of subsamples
     •    Comments relative  to  sample quality
     •    Names of chief scientist  and sampling team
     •    Vessel name.
                                      89

-------
G'-ab Sampl ing—

     Most sampling  of benthic infauna  and  sediments for 301 (h)  monitoring
p-ogr*Tis will be  conducted  using grab  samplers.  Benthic organisms may
be  used  for a variety of purposes,  such  as  1)  comparing species composition
ar.d abundances between potentially impacted  and reference  areas (spatial
trends)  or  between sampling  periods (temporal  trends), and  2)  evaluating
bioaccumulation of  toxic  substances  in  body tissues.   Sediments may be
used  for such purposes as  1)  comparing  characteristics of grain sizes and
concentrations of  conventional  and toxic  substances between  potentially
impacted and reference drees  or between  sampling periods, and 2)  relating
sediment characteristics  to observed distributions of  benthic infauna.
To  be capable of  collecting infauna and  sediments suitable for quantitative
comparisons,  the grab must be appropriate for  the  sediment and depth conditions
in  the study area, and must  be  operated  properly following standardized
procedures.   Holme (1971)  discusses the details of grab selection and oper-
ation.  A thorough reading of that reference is highly recommended.

     Grab Selection—Numerous grab  samplers  are available for collecting
benthic infauna (see Holme 1971).   This variety exists because field conditions
(e.g., weather,  water depth,  sediment  type)  are not uniform, and because
the structure (e.g., species composition,  abundance) and function (e.g.,
trophic relationships, energy flow) of benthic  communities vary spatially
and temporally.  The  grab selected  for use should be able to perform reliably
and efficiently under  the anticipated field conditions and given  the anticipated
types  of benthic communities that  will be sampled.

      To  be  reliable,  a grab must  collect useful samples consistently within
its design  undermost  anticipated  field conditions.   That is, each  time
the grab is  deployed  there should  be  a high probability of collecting  a
quantitative  sample.   Matching sampler design with the predominant  sediment
types is especially  important for achieving a high  degree of  reliability.
The grab should  also  collect representative  samples  of the populations
that  comprise the infauna  (i.e., be efficient).  To be  efficient, the "bite"
into  the sediments should  be deep enough to  collect  most of the organisms
within the jaws  of the  grab.  Moreover, the "bow wake"  associated with

                                      90

-------
trie descending grab should  be as small  as possible  to minimize blowing
away the organisms at the sediment-water interface  immediately before  the
grab contacts  the sediment.

     The installation of mesh  screens instead  of  solid plates on the tops
of the grab jaws will help minimize the "bow wake",  and  will  keep organisms
from  escaping once the  grab jaws are  closed.   The screen mesh apertures
should be the  same as those used  for  sieving samples  in  the field.   The
screens should be hinged  to provide easy access to the surface of the sample
so that grab penetration  can be estimated, the  condition  of the  surface
sediments  can be  observed,  and  sediment subsamples can be collected.  The
exterior of the hinged screens   hould  be  fitted  witn rubber flaps  that
close  when the grab is  being retrieved, but open when the grab is being
deployed.  Although these flaps may  increase  the bow wake  slightly  (even
when they open on descent), they will help minimize  winnowing of the surficial
sediments when the grab is retrieved.

     In  addition to  selecting  the proper kind  of  sampler, it is important
to choose a grab that samples an appropriate area of bottom.   This  should
be done prior  to the major sampling effort using  data from historical studies
or a preliminary survey.  There are two recommended ways to determine  the
appropriate area to sample (Gonor  and Kemp 1978):

     1.   Choose a sample unit size  that  collects a  representative
          complement of the species present, or

     2.   Choose a sample unit size  for which  the variance-to-mean
          ratio for major community variables  is  low.

The  first  criterion is  met by plotting the number of species collected
as a function  of different sample  unit sizes. -At  the  point on  the curve
where the  number  of species approaches  an asymptote,  the  sample unit size
is considered  sufficiently large  to collect  a  representative complement
of  species.  The  second  criterion  is  met by calculating  the variance-to-
mean ratios of various sample  unit sizes  for a given variable.  A  low ratio
is  desirable  for  accurate and precise  estimates of  mean values and for

                                    91

-------
statistical comparisons  of  mean  values among stations.   The variance-to-
r*an ratio  changes  with  sample unit size  because benthic organisms  are
attributed patchily  in  most  cases (Elliott 1971).  When the sample unit
s:ze  is much smaller  than the patch size,  the  ratio generally is low.
As  the  sample unit  size  increases to about  that of the  patch, the ratio
usjally reaches its  highest value.   The ratio then declines with further
increases in sample  unit size.

     In  summary, the  ideal  benthic  grab sampler  is  reliable, efficient,
large enough to collect a  representative complement of species,  *nd  small
erough  to yield  low variance-to-mean ratios  for fiajor community variables
(e.g., to* 1 numbers of species, total  numbers  of individuals). Unfortunately,
S'-ch a  sai pier has yet  to  be  invented. However, consideration of the forgoing
factors  should result in the  selection of  a  dependable  sampler that  yields
information adequate  to  meet the objectives of the monitoring program.
In  most coastal  and estuarine areas, a  0.1-m2 (I.l-ft2)  van  Veen grab and
a 0.1-m2 (I.l-ft2)  Smith-Mclntyre grab are  adequate  in most silty and sandy
substrates.  However, preliminary  tests  should be conducted  for various
samplers in all  areas that  have not been  sampled  previously,  and in areas
wf.ere the sediments  are predominantly clay or  gravel.

     Grab  Deployment  and Retrieval—The grab sampler  should be attached
to  a hydraulically operated cable and should be preceded by a  ball-bearing
swivel   (Figure  3). The  swivel and all shackles connecting  the cable to
the grab should have a  load capacity at least  five  times that  of a  loaded
grab.    It  may be desirable  to keep the  shackles  and  swivel small, however,
as  it may  be necessary for  the shackles to  travel  through  the block at
tne end of the  davit  when  retrieving the  grab.   A snatch block.at  the end
of  the  davit in  Heu of a  standard  block may also  facilitate grab  deployment
and retrieval.   For safety  reasons, a safety pin should be used  to  prevent
the grab f^om  triggering accidentally.  Also,  the  grab should be deployed
with a minimum  of  swinging when  out of the water.   Use  of handling  lines
 (Figure 3)  clipped  to the  hydrowire  is an effective method for minimizing
grab swinging.
                                     92

-------
       DAVIT
                                                SNATCH BLOCK
                         RAIL OF SHIP
Figure 3.   Deployment  of a grab sampler.
                                                    HANDLING UNE
                                                    WITH SNAP HOOKS
                                                        V
                             93

-------
     The grab  sampler should be lowered  and  raised  through the water column
at a  speed of approximately 20 m/min (66 ft/min)  using the vessel's hydraulic
system.   This lowering  rate  generally allows  adequate penetration of the
aev ice into the sediment  while minimizing the chance that  the  sampler  will
be flipped over by  the  weight of excess wire cable  falling  faster than
the parachuting sampler.  If adequate  penetration cannot be  achieved  using
a  controlled descent,  the  sampler can be  allowed to free  fall  the last
1 or 2 m.  However,  extra attention should  be paid to  evaluating  surface
disturbance under  sjch conditions.  Flipping the grab sampler due to excess
caole weight occurs  most  frequently when sampling at depths  in excess of
200  m  (656 ft).  At shallower depth., the weight of the cable is generally
lighter than the weight of  the sampling  device.  When more than 20C i  (656 ft)
of cable is deployed,  care must also  be  taken >.o ensure  that the cable
does not pull itself off  the reel.

     After  the grab has  contacted the bottom, it should be retrieved slowly
at first to permit  the  grab jaws  to  close  properly.   After  the  jaws are
closed,  a retrieval speed  of 20 m/min  (66 ft/min) should be maintained.
When the grab  approaches the water  surface (i.e., when  first sighted),
trie  winch should be stopped to  permit the handling  lines  to be clipped
onto the cable.  The grab can then be raised  slowly, and the handling  lines
can  be used to minimize  swinging  of the grab.  Excessive swinging  is unsafe
and may also compromise the quality of the sample.   When brought on board,
the  grab and  sample should be lowered into a waist-high  stand  designed
to receive it.

     Once  the sampler Is brought aboard,  the hinged screens can be opened
and  the  sample examined.  The sediment should be characterized using the
following qualitative  physical characteristics:   sediment texture;  sediment
color; presence, type,  and strength  of odor; grab  penetration depth [to
nearest 0.5 cm  (0.2 in)];  degree of leakage  (if any);  sediment surface
disturbance; and obvious abnormalities, such as wood debris or large quantities
of mollusc  fragments.   Samples  showing  excessive leakage or disturbance
of the sediment surface should be rejected (Figure 4).   It is also recommended
that samples be rejected if  they  do not achieve the following minimum  pene-
tration depths in various types of sediment:

                                      94

-------
            ACCEPTABLE IF MINIMUM
         PENETRATION REQUIREMENT MET
UNACCEPTABLE (WASHED. ROCK
     CAUGHT IN JAWS)
            UNACCEPTABLE (CANTED
             WITH PARTIAL SAMPLE)
  UNACCEPTABLE (WASHED)
Figure  4.  Examples of acceptable and unacceptable grab ..~..,/ies

-------
     •     Cobble/pebble substrate  cannot be sampled  adequately using
          conventional grab  samplers

     •     4  cm  (1.6 in) for  coarse sand/gravel

     •     5  cm  (2.0 in) for  medium sand

     •     7  cm  (2.7 in) for  fine sand

     •     10 cm (3.9  in) for sandy silt  and  silty sand

     t    10 cm (3.9  in) for si it

     •    10 on (3.9  in) for clay.

See  Shepard (1954,  1963)  for definitions  and  size ranges of the  foregoing
granulometric terms.

     The  acceptable penetration depths are based on the vertical  distri-
bution of organisms  found in these  different sediment types.   Generally,
they are sufficient to  insure that at least  95 percent of the  organisms
and  species that would have been encountered to depths  of  20  cm (7.9 in)
are  captured.   All  of the foregoing  grab sample characteristics should
be recorded on the field  log sheet.

     Log  Sheet—Variables  that should  be recorded on the grab  sampling
log  sheet are:

     •   Geographic location
     •   Date and time
     •   Weather conditions
               Sea state
               Sky state
               Precipitation
                                      96

-------
     •    Station number
     •    Replicate number
     •    Position coordinates
     •    Depth
     •    Sampler description
              Kind
              Area sampled
     t    Number and kinds of subsamples
     t    Sediment characteristics
              Texture
              Color
              P or
              S'ructures
              Debris
              Degree of leakage
              Surface disturbance
     •    Penetration depth
     •    Sieve mesh size
     •    Comments relative to sample quality
     t    Names of chief scientist and sampling team
     •    Vessel name.

Trawl Sampling--

     Most  sampling  of demersal  fishes and  epibenthic  macroinvertebrates
for 301(h) monitoring programs will be conducted using small otter trawls.
Sampled organisms may be used for a variety of  purposes, such  as 1) quantitative
comparison of abundances between potentially impacted and  reference areas
(spatial trends) or between sampling periods (temporal  trends), 2) determination
of prevalences of grossly visible external abnormalities, and 3)  evaluation
of bioaccumulation of contaminants  in tissue  (Tetra Tech 1985a).   To  be
capable of collecting data suitable for quantitative comparisons,  the otter
trawl  must be designed in a  specific manner and operated using  standardized
procedures.  Ideally, at least one member of the trawling team  should  have
ample  experience in using  this device.  Mearns  and Allen (1978)  provide
the most comprehensive description  of how small  otter trawls  should  be

                                      97

-------
cesigned and used for conducting  biological  surveys  in coastal waters.
A thorough reading of that document  is highly recommended.

     Trawl  Design—To  ensi re proper operation, it is essential that the
otter trawl be constructed by an  experienced manufacturer using  a  proven
net  design.  The net should be  large  enough  to capture adequate numbers
of fishes, but small enough to be  handled easily by a crew of two  or  three
persons  on a medium-size [i.e., 9-15 m  (30-50 ft)] vessel. Past experience
with 301(h) data  collection has  shown that  a net with headrope  length of
approximately 7.6 m (25 ft) is an  acceptable size.

     The  footrope of the  Uer  trawl must be weighted to ensure that fishes
living close  to  the bottom  are  adequately sampled.  This capability is
especially critical  for  sampling  flatfishes (i.e., PIeuronectiformes), as
many of these individuals lie partly  buried in bottom sediments when resting.
Mearns  and Allen (1978)  recommend using  a chain rather  than lead weights
on the footrope because its effect on the digging characteristics  of the
net  is easier to  adjust.

     Floats must be  attached to the headrope of  the  otter trawl to hold
the mouth of the  net open when sampling.  When  sampling at depths greater
than 15-30 m (50-100 ft), it is  essential that floats be designed to withstand
high ambient pressure without deforming.  Floats made of styrofoam or cork
tend to  compress under high pressure, whereas hollow plastic floats tend
to burst.

     Otter boards  should be fitted with  a minimum of four towing chains
to allow fine-scaled adjustment  of the angle at  which the  boards  lean when
towed.   A proper angle of  attack  is  critical  to  holding the mouth of the
net  open when  towing.   The  wear  pattern on  the shoe  of the otter board
can  indicate whether a board is  operating properly.  In addition,  the manner
in which the boards spread the net  immediately following submersion also
indicates whether they are functioning correctly.  Mearns and Allen  (1978)
describe how to adjust the otter boards.
                                     98

-------
     The net bridles  should  be  flexible  (e.g., braided nylon)  and  attached
to the hydrowire with a two-  or three-way swivel.  This arrangement minimizes
the  twisting forces on the  net  assembly and  allows it to take the  shape
and orien-ation for which it  was designed.  Mearns and Allen (1978) recommend
that the  length  of the bridles  be approximately three times  the  headrope
length  to ensure that the otter boards  spread properly.

     Trawl Operation--Before  sampling  a transect, it is advisable to slowly
make a  pass along the  isobath of interest  and monitor  its  course with  a
fathometer.   If  the isobath  is  highly  irregular  and requires the vessel
to change course frequently, an alternate transect should  be  considered.
To allow the  otter trawl to  sample a a uniform  depth in a consistent manner,
the transect  should be nearly  straigl1:  or  broadly curved.

     After a sampling transect has been selected, the trawl assembly should
be readied for deployment.   The net  should  be laid out  on the deck  and
scanned for rips.  All lines should be  untangled and their attachment points
checked.  The vessel  should  begin maneuvering toward  the  transect,  and
the  net should  be lowered into the water a  sufficient distance  from the
sampling transect to  ensure  that  it  will  touch bottom at  the beginning
of  the  transect.   Immediately  after  the otter boards enter the water, the
winch should  be stopped and  the boards  allowed to spread.   If the boards
do  not  open  the mouth of the  net properly,  the  assembly should be retrieved
and adjusted. After the otter boards spread  properly,  the trawl assembly
should  be lowered at a slow and steady speed  while the vessel is underway.
If the descent is erratic, or tension is  reduced on  the hydrowire,  the
otter  boards may cross, requiring the  trawl  assembly to be retrieved and
redeployed.

     After the net  touches  bottom, a sufficient length   of hydrowire (i.e.,
scope)  should  be payed out to  ensure that the net is pulled  from a horizontal,
rather  than  vertical position.   If the scope  is  insufficient, the  net will
tend to leave the bottom  and  inadequately sample fishes residing  at the
sediment-water  interface.  In general,  required scope declines with increasing
depth because the additional weight of the hydrowire  enhances the horizontal
                                     99

-------
component of the towing forces.  Based on general  guidance  given by Mearns
anc! Allen (1978), the following scope-to-depth ratios are  reconmended:

     •    5:1 for depths < 20 m (66 ft)

     •    4:1 for depths of 20-75 m (66-246 ft)

     •    3.5:1 for depths of 75-200 m (246-656 ft)

     •    3:1 for depths of 200-350 m (656-1,148 ft)

     •    2:i for depths > 350 m (1,148 ft).

It  is  critical that all stations at  a  particular depth be  sampled using
the same scope. Failure to use a consistent  scope  can result  in  variable
sampling efficiencies among stations.

     After  sufficient scope has been payed out, the winch'should be  locked.
Sampling begins at that point.  When  sampling,  the  vessel  should  proceed
at  a  constant speed of approximately  2.5 kn (4.2 ft/sec)  relative to the
seabed.  If  the speed  is  not held  constant,  the  shape of the net mouth
will  vary and the transect will not  be sampled  consistently throughout
its length.  Gibbs and Mathews (1982) found that an 8-m (26-ft)  otter trawl
fishes most efficiently when towed at about 2.5 kn.  Faster  speeds caused
the footrope to leave the bottom, whereas slower  speeds caused the  spread
of  the trawl to decline and the otter boards and footrope  to  dig excessively
into soft substrates.

     When  a surface current  or substantial wave action  occurs, the vessel
should point into  the  current or  waves when  sampling providing  this can
be  accomplished within  the constraints of  the study plan.   It is  strongly
recommended  that fishing  effort  be  based  on  transect length rather than
time  of trawling.  As shown  in  Figure 5,  a  5-min haul at a  constant vessel
speed  of  2.5 kn (relative to the  water surface)  will cover a distance of
0-385 m  (0-1,262 ft), depending  upon the  speed  of the opposing  surface
                                    100

-------
CO
cc
UJ

UJ
UJ
UJ
              i — i — i — i — i — i
             0.2  0.4  as as  uo  1.2  1.4  is  18 2.0  2.2  2.4 2.6
                       CURRENT SPEED (KNOTS)
   Figure 5.   Transect length as a  function of current
              speed. Net is towed at  2.5 knots for 5
              minutes into an opposing surface current.
                              101

-------
c-rrent.  It is  therefore very  difficult  to sample  a  standardized area
o- the seafloor when fishing effort is  based  on time  of  trawling.

     When a transect has been  covered  completely,  the winch should be engaged.
Sampling ends at that point.   The net  should be  retrieved at a  constant
speed  as the vessel moves forward  slowly.  Constant  tension  should  be maintained
on the net  to  prevent  captured  fishes from  escaping.  Ideally,  the net
assembly should break the water in its fishing orientation.

     After  the net  assembly  is  retrieved  and  secured,  the  catch can be
c-ocessed.  Qualr.y assurance procedures  related  to catch processing are
given  in later  sections of this report concerning bioaccunulatior and  assemblage
cnaracteristics of demersal fishes and epibentnic macroinvertetr-ates.

     Log Sheet—The variables that should be recorded on  the  trawl  sampling
log  sheet are:

      0   Geographic location
      •   Date
      •   Weather conditions
                Sea state
                Sky state
                Precipitation
      •    Station number
      •    Replicate number
      •    Depth
      •    Vessel  heading
       0    Vessel  speed
       •    Surface current
                Speed
                Direction
       •    Hydrowire length
                                       102

-------
     •    Position coordinates and times
               When net begins descent
               When winch Is locked
               When net begins ascent
     •    Sampler description
               Kind
               Headrope length
               Body mesh size
               Cod-end liner mesh size
     0    Number  and  kinds of subsamples
     •    Comments relative to sample quality
     •    Names of chief scientist and sampling team
     •    Vessel  name.

Sample Handling

     After  sample  collection,  proper sample handling ensures that changes
in the constituents of  interest are  minimized and guards against  errors
when  shipping and analyzing samples.  Recommended sample size, containers,
preservation, and storage requirements for each offshore variable  are summarized
in Table 5.   Recommended laboratory methods for measuring receiving-environment
variables are listed  in Table 6.  These requirements and methods  are also
noted  in subsequent sections.  These  requirements  and methods  should be
reviewed in  advance by the laboratory personnel to ensure that sample  sizes,
containers, preservatives,  and all  other specifications are consistent
with laboratory needs and capabilities.

Field Procedures--

     It  is  important  throughout  any sampling and analysis program  to maintain
integrity of the  sample from the  time of collection to  the  point of data
reporting.   Proper  chain-of-custody  procedures  allow the  possession  and
handling of  samples  to  be traced  from collection to  final disposition.
Documents needed  to maintain proper chain-of-custody include:
                                     103

-------
        TABLE  5.   RECOMMENDED SAMPLE SIZES, CONTAINERS, PRESERVATION,
                     AND HOLDING TIMES  FOR OFFSHORE  SAMPLES
, 	 	 	
Measurement
Receiving waterC
PH
Salinity
Temperature
Color
Turbidity
Total suspended solids
Settleable solids
Floating particulates
Dissolved oxygen
Probe
Winkler
Biochemical oxygen
demand
Oil and grease
Nitrogen
Atnmonia-N
Total Kjeldahl-N
Nitrate * Nitrite-N
Phosphorus (total)
Total and fecal
coli form bacteria
Enterococcus bacteria
Chlorophyll a
n i n imum
Sample
Size'

25 mL
200 mL
1 L
SO mL
100 mL
1-4 L«
1 L
5 L
300 mL
300 mL
1.000
1.000
400 mL
500 mL
100 mL
SO mL
250-500
250-500
2-4 L«
Container*

P.G
P.G
P.G
P.G
P.G.
P.G
P.G
P.G
6 bottle &
top only
G bottle &
top only
P.G
6 only
P.G
P.G
P.G
P.G
mL P.G
mL P.G
P.G
Preservative

None
None
None
Cool, 40c
Cool . 40c
Cool . 4<>c
Cool. 40C
None
None
Fix on site; store
in dark
Cool. 4<>c
Cool, 40c
H2S04 to pH<2
Cool , 40C H2SOd to
pH<2
Cool. 40c H2SOa to
pH<2
Cool, 40C H2S04 to
pH<2
Cool, 40c H2SO« to
pH<2
Cool . 40 c
0.0081 Na2S203?
Cool, 40 c
0.0081 Na 282039
Freeze at -20°
Maximum
Holding
Time

Analyze immediately13
Indefinite
Measure immediately*1
48 h
48 h
7 days
48 h
Analyze Immediately"1''
Analyze immediately4
8 h
48 h
28 days
28 days
28 days
28 days
28 days
6 h
6 h*
21 days*
Phytoplankton
1 L
P.G
                         in the dark
                         in a desiccator

                         101 formal in
                                                                        Indefinite"
                                          104

-------
TABLE S. (Continued)

Sediment/ In fauna
Grain size
Total solids
Total volatile solids
Total organic carbon
Biochemical oxygen demand
Chemical oxygen demand
Oil and grease
Su If ides
Total
Water soluble

100 g
SO g
SO g
50 g
SO g
25 g
50 g
50 g
10 g

P.G
P.G
P.G
P.G
P.G
P.G
G only
P,G
P.G

Cool. 40 c
Freeze
Freeze
Freeze
Cool, 40 c
Cool, 40 c
Cool 40 C;
or freeze
Cool , 40 c
zinc acetate
Cool, 40 c
SAOB

6 
-------
TABLE 6.  RECOMMENDED METHODS FOR MEASURING RECEIVING-ENVIRONMENT VARIABLES
Method Reference
Variable
Receiving Water
pH
Salinity
Temperature
Color
Turbidity
Transmissivity
Total suspended solids
Settleable solids
Floating particulates
Dissolved oxygen
Probe
Uinkler
Biochemical oxygen demand
Oil and grease
Nitrogen
Ammonia-N
Total Kjeldahl-N
U.S. EPAa
150.1
-
170.1
110.3
180.1
-
160. 2f
160.5
-
360.1
405.1
413.1
413.2
350.1
350.2
350.3
351.1
351.3
APHAb
423
-
212
2048
214A
-
209Cf
209E
206A9
42 IF
507
503A
503B
41 7A
41 7B
4170
4176
41 7B
4170
41 7E
42 OA
420B
Otherc
ln_ situd
Salinometere
In situd
In situd
-
-
In situd
-
-
-
In situd
SFrTcTTand
and Parsons
(1972)
-
-
In situd

                                    106

-------
TABLE  6.   (Continued)
    Nitrate+nitrite-N

   Phosphorus  (total)

  Total coliform bacteria
  Fecal coliform bacteria
  Enterococcus bacteria
  Chlorophyll £

Phytoplankton

Sediment
  Grain size
  Total solids
  Total volatile solids
  Total organic carbon
  Biochemical oxygen demand
  Chemical oxygen demand
  Oil  and grease
  Sulfides
    Total
    Water soluble
353.2
353.3
365.1
365.2
365.3
418C
418F
42 4C
42 4F
42 4G
908A"
909Ai
908Ch
909C1
                           p.  H4h.j
                           p.  1081.J
                           p.  132" .J
                           p.  1241 ,j
                           U.S. EPAk
                           Strickland
                           and  Parsons
                           (1972)
                           Stofan  and
                           Grant  (1978)
                           Plumb  (1981)
                           Plumb  (1981)
                           Plumb  (1981)
                           Plumb  (1981)
                           Plumb  (1981)
                           Plumb  (1981)
                           Plumb  (1981)

                           Plumb  (1981)
                           Green  and
                           Schnitker
                           (1974)
                                     107

-------
TABLE 6.  (Continued)
  Priority pollutant metals               -              -         Tetra Tech
                                                                   (1986a)
  Priority pollutant organic  compounds    -              -         Tetra Tech
                                                                   (1986a)
  Infauna                                 -              -         Present
                                                                   document
Bioaccumulation (tissue)
  Priority pollutant metals               -              -         Tetra Tech
                                                                   (1986b)
  '   loricy pollutant organic  compounds    -              -         Tetra Tech
                                                                   (1986b)

a Methods recommended in U.S.  EPA (1979b).
b Methods recommended in APHA (1985).
c Methods recommended in sources other than U.S.  EPA (1979b)  or APHA (1985)
when no methods were recommended in the latter two sources.
d This  variable can be measured using an in situ instrument.   The operating
manual for the  instrument should provide all necessary  information  for
proper instrument calibration and measurement of this variable.
« The  instruction  manual for  the salinometer should provide all necessary
information for instrument calibration and salinity determination.
f A  0.40- or 0.45-um membrane filter should be used instead of the glass-
fiber filter recommended in the method.
9 This method is tentatively  recommended by APHA.
n This method can be used whether or not chlorine is present.
i This method can be used only when chlorine is absent.
J Page reference of this method in Bordner et al.  (1978).
k U.S. EPA is currently finalizing a recommended method for this variable.
                                     108

-------
Field  logbook -- All  pertinent  information on field activities
and sampling efforts  should  be  recorded  in  a bound  logbook.
The field supervisor  should  be responsible for  ensuring
that sufficient  detail is recorded in the  logbook.   The
logbook should enable someone  else  to completely reconstruct
the field activity  without relying on the memory of  the
field crew.  All  entires should  be made  in indelible ink,
with each  page signed  and  dated by  the author, and  a  line
drawn  through the unused portions of any page. All corrections
should consist  of permanent  line-out deletions  that  are
initialed by the chief scientist.  At a  minimum, entries
in a logbook should include:

    Date  and time of starting  work

    Names of field supervisor  and  team  members

     Purpose of proposed sampling  effort

     Description  of  sampling  site, including  information
     on  any  photographs that may be taken

     Location of  sampling site

     Details of actual  sampling effort,  particularly  deviations
     from  standard operating procedures

     Field observations

     Field measurements (e.g.,  pH,  temperature,  flow)

     Field laboratory  analytical results

     Sample identification
                           109

-------
         Type and  number of sample bottles collected

         Sample handling,  packaging,  labeling, and  shipping
         information  (including destination).

     Chain-of-custody procedures should  be maintained with the
     field logbook.  While being used  in  the field, the  logbook
     should remain  with the field  team at  all times.  Upon completion
     of the  sampling effort,  the logbook  should  be  kept  in  a
     secure  area.   All  logged  information should be summarized
     and  submitted  to  U.S. EPA after sampling is completed.

•    Sample  labels -- Sample  labels mi'-.t be  waterproof and must
     be securely fastened to the outside and/or placed  inside
     each sample  container (depending  on the  kind  of sample)
     to prevent misidentification  of samples.  Labels must contain
     at least the sample number, preservation technique, date
     and  time of collection,  location of collection, and  signature
     of the  collector.   Labels should  be marked  with  indelible
     ink.  Abbreviated labels may also be placed on  the cap of
     each jar to facilitate  sample identification.

t    Chain-of-custody records -- A chain-of-custody record  (Figure 6)
     must accompany  every  group of samples.   Each  person  who
     has custody  of the sample  must sign the  form  and ensure
     that the samples are  not left unattended unless  secured
     properly.

•    Custody seals -- Custody  seals  (Figure 7) are used  to  detect
     unauthorized tampering  with the  samples.   Sampling personnel
     should attach  seals to all  shipping containers sent  to the
     laboratory  by common carrier.  Gummed paper  seals or custody
     tape should  be  used so that  the seal must be  broken when
     the container holding the samples  is opened.
                                110

-------
SERA
i
i
t
(




(nind Sum R«gl
kgcney SMC
:HAIN OF CUSTODY RECORD
on 10
Sbrth AVWHM
d« WA 98101
PROJECT
LAB *











STATION











DATE











TIME













SAMPLERS: <5«»iw«
SAMPLE TYPE
i
•I











M











»











\











%











i











NUMBt*
Of
CONTAINfNS

i









REMARKS











RELINQUISHED BY: **— RECEIVED BY: a****
RELINQUISHED BY: /SPOT* RECEIVED BY: rspww*
RELINQUISHED BY: 
-------
O.S. ENVIRONMENTAL PROTECTION AGENCY
CLP Sample Management Office
P.O. Bos SIS - Alexandria, Virginia 22313
Phone: 703/337-2*90  - FTS/337-2490
 1.
 2.
 J.
 7. _
 S. .
 9. .
10. .
II. .
I*. .
13. .
1%.
«*•
'*•
17.
'«•
19.
20.
                        SPECIAL ANALYTICAL SERVICE
                               PACKING LIST
                            EPA CUSTODY  SEAL
SA5 Number
Sampling Office:
Sampling Contact:
(name)
(phone)
Sampling DateUh
Date Shipped!
Site Name/Code:

Ship Tot
Attro
Par Lab Use Only
Date Samples RecM:
Received Byi

Sample Sample Oesc. iption Sample Condition on
Numbers Le., Analysis, Matr' Zoncentratioa Receipt at Lab
                                                             For Lab Use Only.

EMYIROHUEMTAL PROTECPOM ACEKCY

MMFUHO.
UQMATUIIB
PNMTMAM8 AH9 nTl«Oi»»eea *a*i*mTi*al*mt
•
I
*
•

IS

    Figure  7.   Examples  of a  sample analysis  request form (above)
                 and a custody  seal  (below).  Equivalent material
                 should  be used for  301(h)  monitoring.
                                      112

-------
For further  information regarding  proper chain-of-custody procedures, consult
the policies  and procedures manual  for the National Enforcement  Investigations
Center (NEIC;  U.S. EPA 1978).

Sample Shipment--

     All preserved  samples should oe  shipped  immediately after  completion
of sampling.  This minimizes  the number of  people handling  samples, and
protects  sample quality and  security.  Guidance for shipping  hazardous
materials  can  be  found  in U.S.  Department  of Transportation (1984).   As
samples are prepared for shipping, the  following  should be kept in mind:

     «    Shipping  ,ontainers  should  be in good shape and capable
          of withstanding rough treatment during  shipping.

     •    Samples should be packed tightly

               Dividers must separate  all  glas-s  containers

               Empty space within shipping  boxes should be  filled so that
               jars  are held securely.

      •    All  containers must be leak-proof.  If a  container  is not
          leak-proof  by  design,  the  interior  should be lined  with
          two heavy-duty  plastic bags and  the tops of bags  should
          be  tied once  samples are inside.   Adequate absorbent material
          should  be placed in the container in a quantity sufficient
          to  absorb  all of  the liquid.

      t    All  samples  should be accompanied by a sample analysis request.
          Variables  to be analyzed by the laboratory,  and total  number
          and kind  of samples  shipped for  analysis should  be  listed
          on  the request  sheet.   An example sample  analysis request
           form  is  illustrated   in Figure  7.   The laboratory  should
           acknowledge receipt  of  shipment  by signing and  dating the
           form, and  returning  a  copy to the designated QA coordinator.

                                     113

-------
    •    A chain-of-custody record for each shipping  container should
         be filled out completely and signed.

    •    The  original chain-of-custody record and  sample analysis
         request should be  protected from  damage  ana placed  inside
         the  shipping box.   A copy of  each should be retained  by
         the shipping party.

    •    The  custody seal  should  be attached  so  that  :he shipping
         box cannot *ie opened  without breaking the  seal.

    f    For shipping containers  carrying liquid samples

              A "This End Up"  label should be attached to each  side to
              ensure that jars are  transported in an upright  position

              A "Fragile-Glass" label  should be  attached  to  the top
              of box to minimize  agitation of samples.

     •     Shipping containers should be sent  by  a carrier'that will
          provide a delivery  receipt.   This will  confirm that  the
          contract  laboratory received the samples and  serve as  a
          backup to the chain-of-custody  record.

     •     All  shipping charges  should  be  prepaid  by the  sender to
          avoid  confusion and  possible  rejection  of package by the
          contract  laboratory.

Laboratory Procedures--

     At  the laboratory, ona  person  should  be  designated custodian  of  all
incoming samples.   An  alternate should also be designated  to serve  in  the
custodian's absence.   The custodian should oversee  the following activities:
                                     114

-------
     •    Reception of samples

     •    Maintenance of chain-of-custody  records

     •    Maintenance of sample-tracking  logs

     •    Distribution of samples for  laboratory analyses

     •    Sending  samples to outside laboratories

     •    Supervision of  labeling,  log  ke<=Ding, data reduction, and
          data transcription

     •    Storage  and security of all  samples, data,  and documents.

     Upon reception of samples, a designated laboratory  person  should fill
out the chain-of-custody record,  indicating  time  and date of  reception,
number  of samples,  and  condition of samples.   All irregularities indicating
that sample security or  quality may have been  jeopardized (e.g.,  evidence
of tampering, loose  lids, cracked jars) should be noted on the sample analysis
request form  and  returned to the  client-designated QA coordinator.-  In
addition, a designated laboratory person should initiate and maintain  the
sample tracking log that will  follow each  sample  through all stages of
laboratory processing and analysis.

     Minimum information required  in  a sample-tracking log includes:

     •    Sample identification  number

      0    Location and  condition of storage

      •    Date and time of each  sample removal and return to storage

      •    Signature of  person  removing and  returning  the  sample
                                     115

-------
     •    Reason for removal from storage

     •    Final disposition of sample.

     All  logbooks, labels, data sheets, tracking  logs, and custody  records
should have proper identification numbers and be accurately filled out.
All  information  should  be written  in  ink.   Corrections should be  made by
drawing a line  through  the error and  entering the  correct  information.
Corrections should  be  signed and dated.  Accuracy  of all data reductions
ard transcriptions should be verified at least  twice.   All samples and
documents should  be  properly stored within the laboratory until the client
a-.-.-horues their  removal.   Security and confidentiality  of all stored material
should be maintained at all  times.   Before releasing analytical results,
all information on sample tags, data sheets,  tracking  logs, and custody
records should be cross-checked to  ensure that data pertaining  to  each
sample are consistent  throughout the record.

     Originals of the  following documents should be sent to the client:

     •    Chain-of-custody  records

     •    Sample-tracking logs

     •    Data report sheets

     0    Quality control records.

Copies of all  forms  should be retained by the laboratory  in  case originals
are lost  in transit.

     When replicate analyses are conducted as QA/QC checks,  it is  recomiended
that they be distributed relatively evenly  throughout  the  full sequence
of samples analyzed,  rather than  being concentrated at  some  point (e.g.,
beginning, end)  of  the sequence.  This  precaution will  enhance the  probability
that if problems develop within part  of a  sequence, they  will  be detected.
                                     116

-------
S-ipboard Laboratory Analyses

     Depending  upon  the size  and capabilities of the survey vessel, many
cf the receiving-environment variables described  in this document can be
analyzed on board.   In  general, the laboratory  procedures  described  in
tms document are  applicable to both shipboard and land-based laboratories.
Tnis  consistency  is  important  to  ensuring that  analytical  results will
te comparable regardless  of which  kind of laboratory generates them.

     Although most laboratory procedures  are similar  between  shipboard
via land-based laboratories, a number  of additional factors must be considered
wnen analyzing  samples at sea.  These  factors relate primarily to the remoteness
cf the shipboard laboratory  from  land-based  support and  th-» movement  and
limited  space of the  survey  vessel.  The major considerations are:

     •    The design  of the  laboratory should be efficient,  with convenient
          equipment locations  and adequate storage  space.

     •    The vessel  should  be  equipped with an uninterruptible power
          supply  that is adequate  for operation of  all  scientific
          instruments.

      t    The laboratory should  be  well-ventilated to  remove  any toxic
          vapors created by  chemicals.

      0    The  temperature  of the  laboratory  should be  controlled,
          especially if variations in  ambient temperature can  influence
          particular analyses.

      •    Adequate lighting  is necessary, especially for  analyses requiring
          color  discrimination (e.g., titration endpoints).   Fluorescent
          lights of the daylight type are recommended (U.S.  Navy  1968).

      •    The laboratory should have adequate water purification apparatus
          or be capable of storing water purified on shore.
                                      117

-------
•    For storing many  kinds of  samples, adequate  refrigeration
     and freezing capabilities are desirable.

t    The laboratory  should never be used as  a general  passageway
     or lounge.

•    The laboratory should be off-limits to unauthorized  personnel.

•    Adequate  safety and first aid equipment should be on board,
     preferably including an overhead quick-pull  safety  shower.

a    Extreme care must be taken when handling  sample   (for quality
     purposes)  and  hazardous  reagents  (for  safety purposes},
     as vessel  movement can sometimes be unpredictable.

•    Backup  supplies and instruments should be on  board so that
     sampling  can continue if a  piece  of equipment  is broken
     or will  not operate properly.  A continuously updated inventory
     tracking system  is useful for maintaining backup equipment.

a    All equipment  should be  properly secured to compensate  for
     predictable and  unpredictable  vessel movements.   Specially
     designed racks are useful for this purpose.

0    Instruments should be checked and  calibrated before sailing
     so that  problems  requiring  land-based  assistance can  be
     solved quickly.

•    Whenever  possible, plastic containers should be used  instead
     of glass because plastic is less susceptible to  breakage.

•    Instruments having digital displays are preferred  over  those
     using analog displays.

a    Pre-printed data sheets  should  be used to ensure that  all
     required information  is recorded.

                                118

-------
 RECEIVING WATER ANALYSES

      QA/QC procedures are presented in  this section for the following 21
 receiving water variables:

      •     pH*
      t     Salinity*
      •     Temperature*
      •     Color
      •     Transparency
      •     Turbidity (transmissivity*)
      •     Total suspended solids
      •     Settleable solids
      •     Floating particulates
      •     Dissolved oxygen (Winkler method)
      •     Dissolved oxygen (Probe method)*
      •     Biochemical oxygen demand (BOO)
      •     Oil and grease
      •     Nitrogen (ammonia)*
      •     Nitrogen (total Kjeldahl)
      •     Nitrogen (nitrate and nitrite)
      t     Phosphorus (total)
      t     Total and fecal coliform bacteria
      •     Enterococcus bacteria
      •     Chlorophyll £
      •     Phytoplankton.

      Samples  to be analyzed for these variables generally will be collected
using water-bottle samplers.  Those variables  followed by  an  asterisk  (*)
may  also be measured using in situ instruments.   Operation of water-bottle
samplers  and in situ instruments are  described  in  the preceding general
methods section.
                                      119

-------
                                                            Receiving Water
                                                                        PH
2H

Field Procedures--

     Col 1ection--Samples  for pH determination should  be collected  in poly-
ethylene or glass bottles with airtight screw caps.   Because  pH  is  unstable
and  cannot be preserved,  these samples should be collected  immediately
after the sampler is  brought on deck.  Only dissolved oxygen  samples should
be  collected before pH  samples.   Because pH of waters not at equilibrium
with the atmosphere  may  change  upon  pxposure  to  the atmosphere, sample
containers should be  completely filled and tightly sealed during collection.

     Prior to sample  collection,  each sample bottle and cap should  be  rinsed
thoroughly with sample water.  This can be achieved by filling the bottle
halfway, sealing  and shaking it, and rinsing the stopper as the wash water
is  discarded.

     A  piece of soft-walled  rubber tubing should be attached to the  outlet
valve of the sampler.  This tubing  should then  be  inserted to the bottom
of the  sample bottle and at least one  full volume allowed to overflow the
bottle.  With the water still  flowing, the tubing should be withdrawn  slowly
from the sample bottle.  Contamination of the sample with air bubbles  should
be avoided.

      After the  tubing has been  removed  from the sample bottle, the stopper
 should be put in place carefully  to avoid  trapping air bubbles.  Once stoppered,
 the sample  should be checked for  bubbles.   If  they are present, the sample
 should  be discarded,  and a new  one taken.   The stopper  on each  accepted
 sample  should be double-checked  to  ensure a  tight seal.

      Process ing--Because pH cannot be preserved, samples  should be analyzed
 immediately after collection.   If a short delay  occurs,  the samples  should
 be  stored  in  the dark at  4° c  and the  storage time  should  be  noted on the
 log sheet.
                                      120

-------
                                                           Receiving Water
                                                                        PH
Laboratory Procedures--

     Analytical  Procedures--Ana1ytica1 procedures are given in U.S.  EPA
Method 150.1 and  APHA Method 423.   As noted  previously,  pH  samples  should
be analyzed as soon  as possible following collection.

     Several  potential  sources of interference with  pH measurements should
be avoided.  Because the response of  the electrode  can  be  impaired  if it
is coated with oily or  iarticuiate material, the electrode should be gently
blotted 01- washed  periodically with a detergent.   Treatment with hydrochloric
acid may be necessary to remove some kinds of film.   Temperature can influence
pH measurements by altering electrode output and  changing  the pH  inherent
in the sample.  The  first source of temperature interference can be controlled
by using a pH meter  with  temperature compensation  or by  calibrating  the
meter at the temperature of the samples.  Because the second kind of temperature
interference cannot be controlled, the temperature at which the pH  determination
of each  sample  is  made  should be  logged  and reported.   It  is recommended
that  sample  temperatures  never differ by more than 2° C from that of the
buffer solution.

     When  pH measurements are being made, it is critical  that the sample
be stirred at a constant rate  to provide drift-free (<0.1 pH  units) measure-
ments.   The  rate of stirring should  minimize air transfer at the  surface
of the sample.   At least 30  sec  should be  allowed for each measurement
to stabilize.

      Calibration and Preventive Maintenance--Calibration procedures for
the pH meter  should follow specifications  of the manufacturer.   General
guidelines are given by U.S.  EPA  (1979b)  and  APHA (1985).

      Buffer  salts  can be  purchased  as a  solution that has been  calibrated
 to National  Bureau  of Standards salts.   The  pH meter  should be  calibrated
at a  minimum of  two points  that bracket the  expected  pH value of the samples
                                     121

-------
                                                           Receiving Water
                                                                       pH
and  that are three or more pH units  apart  (U.S. EPA 1979b).   Prepare fresh
buffer solutions at  least  every month  to  avoid erroneous calibration as
a result of mold growth or contamination.

     Preventive maintenance  procedures  should  follow specifications given
by the manufacturer  of the pH meter.   In general, verification of electrode
performance and meter performance is  the  only operator service recommended.
An electrode should  be replaced when  it no longer meets  span requirement
and does not improve with rejuvenating procedures.

     Qua! ity Control Checks—The pH meter  should be calibrated  at the beginning
of each series of samples and after each group of 10 successive measurements.
It  is  recommended  that duplicate pH  determinations be made on at  least
10 percent  of  the total  number of samples.  As an  independent  check,  a
U.S. EPA reference sample  should be analyzed at a minimum of  every 3 mo.

     Corrective Action—If  the  pH  meter  does not  appear  to be operating
correctly,  consult the manufacturer's troubleshooting  guide.  Some common
problems  include a dirty electrode, failure to fill  the reference portion
of the electrode with  internal  solution, and  inadequate stirring.

      Data  Quality and  Report ing--A precision of +_ 0.02  pH unit and  an accuracy
of 0.05 pH unit can be achieved  under the  best circumstances.   However,
the  limit  of accuracy  under most circumstances is ^0.1 pH unit  (APHA 1985).
A precision of  0.1  pH unit is considered acceptable (U.S. EPA 1979b).

      Measurements  of pH  are reliable only when the instrument has been
 calibrated by  standard buffers bracketing  the desired range.   Samples having
 a pH  greater   than 10 may require  a special  probe to correct for "sodium"
 error.  However, pH values as high as 10  are not likely to  be  encountered
 in most coastal waters.

       It is recommended  that pH values be reported  to the nearest 0.1  unit.
 In addition, the ambient  temperature  at  the time of measurement of each

                                     122

-------
                                                            Receiving Water
                                                                         PH

sample should be reported  to the nearest degree C.  Results of all  determina-
tions should be reported,  including QA replicates.   Any  factors that may
have influenced sample  quality should also be reported.
                                     123

-------
                                                           Receiving Water
                                                                 Salinity
Salinity

Field Procedures--

     Col1ection--Two kinds of bottles are acceptable for collecting salinity
samples.  The first is made of borosilicate glass and has an airtight septum
stopper.  The second is made of low-pressure polyethylene and has polyethylene
inserts in the screw caps.  Soft glass bottles,  high-pressure  polyethylene
bottles, and ground glass stoppers snould not be used.

     Prior  to sampling, the collection  bottles  should be "seasoned"  by
filling them with  seawater.  Bottles should remain  upside down  in the case
until  the sample  is taken.  Bottles with chipped edges or loose caps should
not be used.

     Each1 collection bottle and  cap  should be rinsed at least three  times
with sample water  before  the sample is  collected.   No salt  crystals should
remain on the bottle  or stopper.  The bottle and stopper  should  not  be
contaminated by contact with any  surface.   If contamination occurs, the
rinsing step should  be  repeated.

     After the bottle and  stopper have been  rinsed thoroughly, the  bottle
should  be  filled  to  approximately 90  percent of the bottle volume with
sample and  sealed.  The stopper should  be  double-checked for a tight fit.
The  external  label on each bottle  should be filled  out  completely.

     Processing—No reagents are necessary  to preserve  the  salinity samples.
Bottles should be stored upright  after samples have  been collected.   If
necessary,  properly  sealed salinity  samples can  be stored indefinitely
before analysis (Strickland  and  Parsons 1972).
                                     124

-------
                                                          Receiving Water
                                                                 Salinity
liooratory Procedures--
     Analytical Procedures--The salinometer  manufacturer should provide
a detailed description of how  to  use the instrument.   To avoid  heating
t^e  sample bottle, it should  oe  propped up rather  than  gripped Dy hand
d-jring the salinity determination.  The sample cell  bowl  should be  filled
slowly  to avoid introducing  bubbles.  The cell  should be rinsed thoroughly
with distilled  water between successive samples.

     Calibration  and  Preventive Maintenance--Calibration prod lures should
follow the specifications given by the manufacturer of the  salinometer.
It  is  recommended that the  primary standard be  Copenhagen seawater and
tnat all secondary  standards be  based on this primary standard.  Secondary
standards should  consist of  filtered  seawater collected  from the open ocean
at a depth of at least 50 m (APHA 1985) and should be periodically  checked
against the primary  standard  to  guard  against  contamination or drift.
The  secondary standards should be equilibrated  to  the temperature of the
samples before calibration  begins.  Secondary standards  should be stored
 in glass containers and protected from  evaporation and light.

     Preventive maintenance procedures  should follow the  salinometer manufac-
turer's recommendations.   These include  periodic  cleaning of the sample
cell bowl, greasing the threads on the water-trap jar, tightening  all water
connections, checking  the temperature circuit calibration,  and lubricating
 tne pump, pump motor, and stirrer motor.  It is critical that the  sample
cell bowl be  kept clean.  Normally, the bowl  should  be cleaned daily.
 However,  if the sample water  is  very dirty, hourly cleaning  may be necessary.

      Quality Control  Checks—Two standard sample determinations  should
 be made before the start of each  series of samples.   In  addition, one standard
 sample should be analyzed  after each group  of 10  successive samples,  to
 monitor instrument drift.  It is recommended  that duplicate determinations
 be made for  at least  10 percent of the samples  analyzed.
                                     125

-------
                                                           Receiving Water
                                                                 Salinity
     Corrective  Action—If the salinometer does not appear to be operating
properly, the manufacturer's  troubleshooting  guide should  be consulted.
Several  common problems  include failure of  the  null indicator to  show a
deflection, failure  of the thermometer circuit, excessive  salinity  balance
drift,  inability to fill  the sample cell  completely, and failure  of the
stirrer to operate properly.

     Data Quality and Reporting — It is recommended that  salinity determinations
be made using  an  induction salinometer.   Other,  less common procedures
include the hydrometric  and  argentometric  methods  see Methods  210B and
210C Of APHA 1985).

     A  precision of +0.1 ppt  is possible using  an induction salinometer.
Conductivity measurements should be converted to salinity values using
standard tables corrected  for  temperature.  If possible,  salinity concentrations
should  be reported  in ppt  to the nearest 0.01  unit.  Results of all deter-
minations should be reported,  including QA replicates and  standards.  Any
factors that may have  influenced  sample quality should  also be  reported.
                                      126

-------
                                                           Receiving Water
                                                               Temperature
Temperature

Field Procedures--

     Col 1 ection--Temperature can be measured  using  a mercury-filled Celsius
thermometer on  samoles collected in glass or  plastic containers.   The  ther-
mometer should have a scale etched on capillary glass  for 0.1° C  increments
and a minimal  thermal capacity to permit rapid equi1 ibration.  Temperature
can  be  measured -in situ  using a reversing  thermometer  or a thermistor.
Of these two _i£ situ  instruments,  the then-istor is more accurate, but
also more  expensive.

     Processing—Because temperature cen change rapidly  after a  sample
is removed  from ambient conditions, temperature determinations by thermometer
should be  made immediately after sample collection.

Laboratory Procedures—

     Analytlcal  Procedures—Methods  for making  temperature measurements
are described in U.S.  EPA  Method  170.1 and in APHA Method 212.   It  is critical
that the  measuring device be  adequately immersed  in the sample  and  allowed
to completely equilibrate  (i.e., the temperature reading stabilizes)  before
temperature is determined.

     Calibration and Preventive Maintenance—Each kind  of temperature-measuring
 instrument should be calibrated  frequently  against a  National   Bureau  of
 Standards (NBS)  certified thermometer that  is used with its certificate
 and  correction chart.  An  NBS thermometer  is recommended because some  commercial
 thermometers may be as much  as  30  C  in error  (APHA  1985).

     To prevent breakage,  it is recommended that  each  thermometer  be enclosed
 in  a metal case.  If a mercury  thermometer  is broken,  samples or containers
 in  the  vicinity of the exposed  area  may be contaminated  by the mercury.
                                     127

-------
                                                          Receiving Water
                                                              Temperature
     Quality  Control  Checks—Each temperature-measuring  instrument should
be calibrated against a N8S standard  thermometer at  least every week.
It is  recommended that calibration  be conducted  daily when a temperature
violation is suspected.

     Corrective  Action--If  the temperature-measuring instrument cannot
be calibrated consistently against the  N8S thermometer,  it should be repaired
or replaced.

     Data Quality and Reporting--Pre. ision and accuracy have  not been determined
for temperature measurements  (U.S.  EPA  1979b).   If possible, temperature
measurements should be reported  to  the nearest  O.Olo C.   Results of all
determinations should be reported, including  QA replicates and  standard
checks.   Any factors that may have  influenced sample quality should also
be reported.
                                     128

-------
                                                          Receiving Water
                                                                    Color
Color

Field Procedures—

     Col lection—Color  samples shoula be collected from  the water sampler
In glass or  plastic bottles.   Bottles should be  cleaned thoroughly  before
use.   Because  biological  activity may change color characteristics, samples
should be coMected and analyzea  as soon as  possible after the  sampler
is Drought  on  board.   It is recommended that  color samples be collected
immediately  after dissolved r tygen  and pH  samples.

     Processing—Because no reagents  can be added to preserve color, samples
should be analyzed as soon  as  possible after collection.  If a delay occurs,
samples  should be held at 4o c for  no more than 24 h.  The length of delay
should be recorded on the log  sheet.

Laboratory Procedures—

     Analytical  Procedures—Color should be determined using the spectrophoto-
metric methods  described in U.S. EPA  Method 110.3 and in APHA Method 204B.

     The recommended method of color  analysis is very sensitive to turbidity.
However, the optimum filter  media to remove  turbidity without  removing
color  has not  been found.   Recommended  methods include filtration  through
a 0.45-um filter and centrifugation.  The spectrophotometric method is
also sensitive  to sample pH.

     Calibration  and  Preventive Maintenance—For  the spectrophotometric
method, the instrument should  be  set to read 100-percent transmittance
on a  distilled water blank.   All  determinations should be  made  within  a
narrow spectral  band.

     Preventive maintenance procedures  for  the  spectrophotometer shoula
follow the manufacturer's recommendations.

                                    129

-------
                                                          Receiving Water
                                                                   Color
     Quality Control  Checks — It  is  recommended that duplicate color deter-
minations be made on at least 10 percent of the total samples analyzed.

     Corrective Action — If the  spectrophotometer does not appear  to be
operating properly, the  manufacturer's troubleshooting guide should be
consulted.

     Data Quality  and Reporting—A:  present, precision and accuracy data
are  not  avail  t'.e for the  spectrophotometric method  (U.S.- EPA 1979b) .
"esults  of th'S method  should be reported  at pH 7.6  and at the origina"1
pH in terms of  dominant wavelength hue  (to  the nearest 1.0 nm), luminance
(to nearest 0.1 percent), and purity  (to  nearest 1.0 percent).  In addition,
the kind of spectrophotometer, number of  selected ordinates (10 or  30),
and  the  spectral  band width  (nm) should  also be reported.  Results of all
determinations  should  be  reported, including  QA replicates.  Any factors
that may have influenced  sample quality should  also be  reported.
                                     130

-------
                                                            Receiving  Water
                                                               Transparency
Transparency
     Transparency is  measured  using a Secchi disk.  This device is  a circular
plate with a standard diameter  of  30 cm  (U.S.  Navy 1968).   The  top side
of  the  disk should  be  white.  A  ring  attached  at  the center  of the disk
allows a graduated line  to be  secured.  A 2-4 kg weight should be  attached
centrally to the  underside of  the disk to ensure that the device  will  sink
rapidly and vertically.   The  deployment line  should be made of  material
that will not stretch substantially after repeated use (e.g., braided dacron).

     Transparency measurements should be made by Ijwering the  Sec;hi disk
from the shaded side of the survey  vessel until the disk is barely perceptible.
This depth should be recorded to  the  nearest  0.5  m and the  disk  should
continue to be  lowered  until  it  is no longer  visible.   The disk should
then be raised  slowly  until  it is again barely visible..  This second depth
should also be  recorded.  The  average of the  two  depth  readings (i.e.,
downward and upward)  should be reported to the nearest 0.5 m as the measured
transparency value.

     Because Secchi disk readings are dependent upon the available illumination,
they vary with time of day, cloud formation, and cloud cover.  Secchi disk
readings also vary  with the observer because of interpersonal  differences
in visual ability.  Thus,  to standardize these readings, repeated measurements
should  be made by one  Individual under similar conditions of illumination.
Because these criteria are not always achievable, associated meteorological
data at the time of measurement  and the name of the person making the  deter-
minations should be included on  the log sheet with the Secchi disk  readings.
                                     131

-------
                                                           Receiving Water
                                                                Turbidity

Turbidity

Field Procedures —

     Col lection—Turbidity samples  can  be collected  in glass  or  plastic
containers.   Samples should be removed from  the sampler as soon as possible
after  collection to minimize settling  of suspended material  within  the
sampler.  Sample containers and lids  should  be rinsed thoroughly with sample
water  before samples are  collected.  Turbidity can be estimated  in  situ
as transmissivity using a transmissometer.   A discussion of such  . • situ
instrumentation is  provided earlier in the general methods section.

     Processing—Because turbidity  samples cannot be preserved adequately,
they should be analyzed as soon as possible  after collection.  If  a delay
occurs, samples should  be held at  40 C for no  more than 48 h to minimize
microbiological decomposition of  solids.  The length  of delay  should be
noted  on the log sheet.

Laboratory Procedures—

     Analytical Procedures—The nephelometric method  is  described in U.S. EPA
Method 180.1 and in APHA  Method 214A.  For turbidities greater than  40  NTU,
samples should  be  diluted with one or more volumes  of  turbidity-free water
until  the turbidity falls below 40 NTU.  Turbidity-free  water  is distilled
water  passed  through  a  membrane filter with a 0.2-um pore size.  Samples
 should be  shaken well  to thoroughly disperse  solids,  and resulting  air
bubbles should be allowed to  dissipate before the sample is analyzed.

      Interference with  turbidity measurements arises from several  sources.
 Because the  presence of floating  debris  and  coarse  sediments that settle
 out rapidly will  give low  readings, readings  should  be  made as soon as
 possible after  sample agitation.   Finely divided air bubbles will affect
 results  in  a  positive direction.   If present, the bubbles should be allowed
 to dissipate before a reading is made.  Finally, dissolved  substances that

                                      132

-------
                                                          Receiving Water
                                                                Turbidity
absorb  light  in  the sample will  reduce  turbidity  readings.  Any color in
the sample should therefore  be noted.

     Calibration and Preventive Maintenance—It is  recommended that a standard
suspension of  formazin be used to  calibrate  the  nephelometer.   Formazin
provides  a more  reproducible turbidity  standard  than do  other materials
used in the past.  The formazin  standard suspension  should be  prepared
daily (APHA 1985).  Commercially available standards  such as  styrene divinyl-
benzene beads  (trade name AMCO-AEPA-1), can  be substituted  for  formazin
if  they ar«- demonstrated to be  equivalent to  freshly prepared formazin
(APHA 1985).  Standards measured on  the nephelometer  should  cover  the rar.c-e
expected  for the samples.  At least  one standard  should be run  in each
instrument range  to be used.   The  instrument  should give stable  readings
in  all  sensitivity ranges used.   If a precal ibrated scale  is  not  supplied,
calibration curves should be prepared for each instrument  range to  be used.

     Quality  Control  Checks — The  nephelometer should  be calibrated at the
start of each series of  analyses  and after  each group of 10  successive
samples.   Duplicate analyses should  be conducted  on at  least  10  percent
of  the  total number of samples.

     Corrective  Action—If the nephelometer will  not stabilize  in any of
the relevant ranges  or if the instrument does  not appear to  be functioning
properly  in any other aspect,  the manufacturer's troubleshooting  guide
should  be  consulted.  Sample tubes  that  become scratched or  etched should
be  replaced.

     Data  Quality and Reporting—Because the  nephelometric method of turbidity
measurement is more  sensitive than the  visual  method, the former  is recom-
mended.  Limited  precision data indicate that standard deviations of measure-
ments  vary directly  with  the level  of turbidity  (U.S. EPA 1979b).  Accuracy
data are  not available  at present.   The  sensitivity of the nephelometer
 should allow detection of a turbidity  difference of 0.02  unit or  less  in
waters with  turbidities less  than 1.0  unit.   Results should be reported

                                     133

-------
                                                           Receiving Water
                                                                 Turbidity
in nephelometric turbidity  units (NTU).   U.S.  EPA  Method 180.1 describes
the nearest reporting  units as a function of the raige of values measured.
Results  of all determinations should be reported,  including QA replicates.
Any factors that may have influenced sample quality  should  also be reported.
                                      134

-------
                                                            Receiving  Water
                                                     Total  Suspended Solid-?
Total Suspended  Solids (TSS)

Field Procedures—

     Collection—Samples should  be collected in glass or plastic  bottles.
Samples should  be  collected soon after the  sampler is brought on  board
to minimize settling  of suspended material within the sampler.   Nonrepresen-
tative particulates such  as leaves  and  sticks  should be  noted, and  then
excluded from the sample.

     Processing—Total suspended solids samples cannot be preserved adequately
and  should therefore  be analyzed as  soon as possible after collection  (APHA
1985).  If a delay occurs, samples should be held at 40 C to minimize micro-
biological decomposition  of  solids.  The  length of delay  should not exceed
7 days and should be  noted  on the  log sheet.

Laboratory Procedures—

     Analytical Procedures—Suspended  solids determination should be made
according  to procedures described  in U.S. EPA  Method 160.2  and APHA Method
209C.   However, a 0.40-  or  0.45-um  membrane filter  should  be used to  remove
suspended  solids instead  of  the  glass fiber filter specified in the U.S. EPA
and  APHA methods.  These  filters are the  ones  used most commonly, for oceano-
graphic work  in coastal waters.

      The  drying temperature of  the  filtered residue  can  influence results
because temperature and time of  heating affect  weight losses due  to volatili-
 zation of  organic matter, mechanically occluded water, water  of crystallization,
 and  gasses from heat-induced  chemical  decomposition,  as well  as weight
gains due  to  oxidation (APHA 1985).   Thus,  drying temperature must be carefully
controlled and  not allowed to deviate  from the reccrrmended range of 103-1050 C.

      To avoid  contamination, filters should  be handled  with  forceps  during
 all  steps  from  initial  to final weight determinations.  Filters  should

                                      135

-------
                                                           Receiving  Water
                                                    Total  Suspended  So>ids
a" ways  be stored in  a desiccator when cooling.  When filtering the samples,
it is critical  that  the filter is  seated  tightly on the  surface of  the
filtration apparatus and  that all  holes  in  the crucible are covered.   To
ensure complete removal  of  salts after  filtering the sample,  the filter
snould be rinsed with a  minimum of three successive 20-nt portions of distilled
water.   It is  recommended  that enough  water be filtered  to ensure that
at  least 5 mg  of  residue  is collected.   Because excessive residue  on  the
filter rjiay form a water-entrapping crust, the sample size should be limited
to  that which  yields  less than 200  mg of residue (APHA 1985).  Prolonged
filtration times resulting   -on filter clogging may produce high results
due  to  excessive solids cap':ure on  the clogged filter.  Therefore filtering
should be terminated  before  any evidence of clogging is noted.

      Calibration and Preventive Maintenance—The  analytical balance should
be calibrated weekly  using  standard  weights according to the manufacturer's
instructions.   It is  recommended  that theibalance have a minimum accuracy
of 0.1 mg.  The manufacturer's preventive maintenance  procedures  should
be  followed carefully.

      Quality Control Checks—For  each weight determination, filters should
be  run  repeatedly through the drying/cooling  cycle until  the weight loss
 is  less than  4 percent of the previous weight or 0.5 mg, whichever is  less
 (APHA 1985).   Duplicate analyses  should be conducted on at least  10 percent
of  the  total  number of samples.   A filter blank should  be  taken through
the  preparation, drying, and desiccation steps  for each batch  of samples
to  monitor changes in filter tare weight.  U.S. EPA reference samples  should
be  analyzed quarterly to check the  overall accuracy of  the method.

      Corrective Action—If  the analytical balance will  not  produce repeatable
measurements  within  0.1 ng, the manufacturer's troubleshooting  guide  should
 be consulted.  If the  filter becomes  clogged during  filtration, it  should
 be discarded  and the analysis should  be  repeated  using  a  clean filter.
 To prevent  clogging of the second  filter,  the  volume of sample analyzed
 should  be reduced.

                                      136

-------
                                                          Receiving  Water
                                                   Total Suspended  Solids
     If the weight of the  filter  blank is not consistent after drying  and
cooling, the desiccant should be checked.   A color-indicating desiccant
is recommended,  so that spent desiccant  is easily detected.  Also,  the
seal  on the desiccator should be checked and,  if necessary,  ground glass
rims should oe greased or "0" rings replaced.   Cooling time should be closely
monitored,  so weighing times are consistent between batches of samples,
thereby minimizing the need  for multiple weighings.

     Data Quality and Reporting—Precision of res- its varies directly with
the concentration of suspended matter and, at lo/  levels,  the ratio  of
the weight of  the suspended matter  to the weight of  the  filter.  There
are no procedures for determining  the accuracy of field  measurements  of
suspended  matter.   Total  suspended solids measurements should be reported
as mg/L to a minimum of two significant figures.   Results of all determinations
should  be  reported  including QA replicates and filter blanks.  Any factors
that may have influenced  sample quality should also be reported.
                                    137

-------
                                                           Receiving Water
                                                         Settleable Solids
Settleable Solids
Field Procedures—

     Col 1 ect ion—Settleable solids samples  should be collected  in  1-L glass
or plastic containers.   The sample container selected for use  should be
cnecked to ensure that material in suspension  does not adhere to the container
walls.  Samples should be collected soon  after the sampler is brought on
board,  to minimize  settlement of suspended material within  the sampler.
Nonreprf ;ent?tive particulates such as  leaves and sticks  should  be noted
and then  excluded from the  sample.

     Process ing--Settleable  solids  samples  cannot be preserved  adequately
and should therefore be  analyzed as soon as  possible after collection (APHA
1985).   If a delay  occurs,  samples should be held  at 40 C to minimize micro-
biological decomposition of solids.  The length of delay should  not  exceed
48 h and  should be  noted on the log sheet.

Laboratory Procedures—

     Analytical Procedures—Procedures used  to determine settleable solids
concentrations  are presented in U.S.  EPA Method  160.5  and in APHA Method
209E.   The sample  should  be well-mixed before introduction to the Imhoff
cone.   It is critical that  floating material is not  included with settleable
material.

      Calibration and Preventive Maintenance—Not applicable.

      Quality Control Checks--Duplicate analyses should be  conducted on
 at  least 10  percent of the  total  number of  samples.

      Corrective Actions—If suspended  material is found  to adhere to the
 sides of the  sample collection containers, a different type of  container
 should be used.

                                     138

-------
                                                            Receiving Water
                                                          Settleable Solids

     Gata  Quality and Reporting — The practical lower limit of measurement
depends on sample composition and generally is in the range of 0.1 to 1.0 mL/L
(APHA  1985).  Precision  and  accuracy data are not available at present.
Settleable solids  concentrations should  be reported as mL/L to a minimum
of two significant figures.  Results should be reported  for all determinations,
including QA replicates.  Any factors that may have influenced sample quality
should also be reported.
                                     139

-------
                                                           Receiving Water
                                                     Floating Participates
F". aati-iq Particulates

Field Procedures—

     Col lee tion—A minimum of 5 L of sample  snould be collected  in a glass
or plastic container.  The container should  be rinsea thoroughly with sample
water before sample collection.

     Processing — For comparable results,  samples must be treated  uniformly
throughout sampling and  handling.  Analyses  should be  conduct-d as  soon
as possible after sample collection.

Laboratory Procedures--

     Analytical  Procedures—Floating particulates  should be analyzed  in
accordance with APHA Method 206A.  At present, this method  is tentatively
recommended by APHA.   Because even slight differences in  sampling and handling
can give large differences  in measureable floating particulates, all  samples
should  be treated  uniformly, preferably  by adequately mixing them before
flotation.  When mixing the sample, care should be taken  to avoid  extensive
air  entrapment through  formation of  a  large vortex.   Because temperature
variations can affect results, all tests should be conducted at a constant
temperature, preferably 20° C.

     Calibration and Preventive Maintenance—No  calibration procedures
apply to the recommended method.  To prevent oil and grease from sticking
to  the  analytical  equipment, all  internal  surfaces  should be coated  with
TFE.

     Quality Control Checks —It is recommended  that duplicate  analyses
be conducted  on  a  minimum of 5  percent of  the total  number of samples,
with an additional  5 percent of  the samples checked  for recovery.
                                     140

-------
                                                           Receiving Water
                                                      Floating Participates
     Corrective Actions—If oil  and  grease  appear  to  be sticking to surfaces,
TFE coatings  should be renewed.   If  recovery  jrops below 90  percent, samples
should be reanalyzed and each step of the analysis carefully scrutinized.

     Data Quality  and  Reporting—The minimum detectable  concentration of
floating particulates using the recommended  method  is 1 mg/L.   Precision
varies with  the concentration of suspended  matter  in the sample.  A  coefficient
of variation  of 5.7 percent has been achieved using five replicate  samples
(APHA  1985). Although there  is  no completely  satisfactory procedure  for
determining thp iccuracy of the method, approximate  recovery can be  determined
oy running  a .econd test for floatables on  all water discharged  throughout
the analytical procedures, except  for the last 10 ml.  Typical recoveries
exceed  90 percent (APHA  1985).   Concentrations of floating  particulates
should be reported  in mg/L to the nearest 0.1 unit.  Results of  all determi-
nations should  be reported,  including QA  replicates and  recovery checks.
Any factors  that may have  influenced sample quality shou.ld also  be reported.
                                     141

-------
                                                           Receiving  Water
                                                 Dissolved Oxygen (Winkler)
Dissolved Oxygen (Winkler Method)
Field Procedures--
     Col1ection--Prior to  sample  collection,  the fixing reagents should
be prepared and the dispensing aoparatus  should be  filled.  The accuracy
of the volumes being dispensed should be checked  and no air should be trapped
in the system.  It is  recc-mended that 300-mL glass  BOO bottles with  ground
glass stoppers be used for the Winkler method.

     Oxygen  samples must be  :he  first ones collected from the sampler  and
they should be collected immediately after  the  sampler is brought on  board.
It  is  recommended  that a  piece of  soft-walled rubber tubing be connected
to the discharge valve of the sampler to prevent air bubbles from contaminating
the  sample during  collection.   The tubing should be .soaked  in seawater
prior to use to prevent air bubbles  from collecting  inside.

     After  being attached  to the  sampler, the tubing  should  be  flushed
with sample water to remove air bubbles.   The  sample bottle and  stopper
should  then be  rinsed  thoroughly with sample water.  After rinsing,  the
tubing should be inserted to the bottom of  the  sampling bottle.  The  bottle
should  be filled  slowly until  at least half full, and then filled  rapidly
thereafter.  At  least one  full  bottle volume  of  sample  should overflow
the bottle before the  tubing is removed.  After the  tubing  is removed slowly,
the stopper should be  carefully put in place with a twisting motion while
water  is displaced from  the  bottle.   Once stoppered, the  sample should
be checked  for  air bubbles.  If bubbles  are present,  the sample  should
be  discarded and  a new sample collected.  Acceptable samples should be
fixed as soon as possible after collection.

     Processing—The stopper  should be  carefully removed from the bottle
without  agitating the  sample.  Each fixing reagent should be added by gently
placing the  tip of the pi pet slightly below the surface  of the sample and
gently pushing the plunger.  The plunger should  not be  released  until  the

                                     142

-------
                                                           Receiving Water
                                                 Dissolved Oxygen (Winkler)
pi pet  has been removed from  the sample.   The pipet tip should be rinsed
with distilled water before being returned  to the  reagent bottle.

     After the fixing reagents  have been  added,  the bottle should be carefully
stoppered without  introducing air bubbles.   Excess fluid around the outside
of the stopper should be poured off and the sample bottle should be inverted
5-10 times to thoroughly disperse the  precipitate.  It  is also good practice
to  invert bottles several  times approximately  20 min after fixation to
ensure thorough dispersion of the precipitate.

     After allowing  the precipitite  to settle  for 10-15 min, the stopper
should be removed  and sulfuric  acid should be added to the  sample  in  the
same manner  as the fixing  reagents.  The  stopper should then be replaced
and the bottle inverted  until  all of the  precipitate has  dissolved.   If
the precipitate fails to dissolve,  it should be allowed to settle again
and additional  sulfuric acid  should be  added  to  the sample.   It is critical
that all  of  the precipitate be dissolved  before  samples are stored.   Also,
it is critical  that samples not be allowed  to stand longer  than 8 h  before
sulfuric acid is added, as  erroneous measurements may result.

     Preserved  dissolved  oxygen  samples  should be  stored in the dark at
10-200  c.  Samples should  be analyzed  as soon as  possible after collection,
and storage time should not exceed 8 h.  The length of storage should  be
recorded on the log sheet.

Laboratory Procedures--

     Analytical Procedures—The  recommended modified  Winkler method  for
saline water  is described in  detail  in  Strickland  and Parsons (1972).

     Calibration  and  Preventive Maintenance—Methods of standardizing  the
thiosulfate solution are presented by Strickland and  Parsons (1972).   It
is recommended that one person perform the standard  and sample titrations
because of subjectivity in  the  color of the endpoint.

                                   143

-------
                                                           Receiving  Water
                                                Dissolved Oxygen (Winkler)
     Preventive maintenance  is limited to ensuring that reagent disoensing
and titrating equipment  is clean and functions  properly.

     Quality Control Checks—All  standard titrations should be duplicated.
It is  recommended that duplicate analyses be conducted on at least 10 percent
of  the  -otal samples.   Four  replicate  reagent blanks should be run once
a week during a cruise,  or whenever a reagent is changed.

     Corrective Action--!  the results obtained by  running duplicate standard
titrations of the thiosulfate  solution  do not  agree within ^0.05 ml,  the
titrations  should  be  repeated until  agreement  is achieved.  All reagent
dispensers should be checked for  bubbles, and the amounts of reagents delivered
should be verified.

     Data  Quality and  Reporting—Using the modified Winkler method, repro-
ducibility for field samples is approximately 0.2 mg/L  of dissolved  oxygen
at  the  7.5  mg/L  level  (U.S.  EPA  1979b).  Duplicate  titrations made during
standardization of reagents should agree within  ^0.05 ml.  With  careful
collection  and  treatment of samples, dissolved oxygen as  low as 1 percent
of  saturation can be measured.   Dissolved oxygen  concentrations  should
be  reported in  mg/L to the  nearest 0.1 unit.  Results  should be reported
for all determinations, including  QA replicates and  reagent blanks.  Any
factors that may have influenced sample quality  should  also  be reported.
                                     144

-------
                                                           Receiving  Water
                                                  Dissolved Oxygen  (Probe)
Dissolved Oxygen  (Probe Method)

Field Procedures--

     Conection--0xygen  samples should  be the first  ones collected  from
the sampler and they should be  collected  immediately after  the  sampler
is brought on board.  It is recommended  that a piece of soft-walled rubber
tubing be connected to the aischarge valve  of  the  sampler  to  prevent air
bubbles  from contaminating the sample during collection.  The  tubing should
be soaked in seawater prior to use to prevent air  bu». >les from  collecting
inside.

     After  being attached to  the sampler, the plastic  or rubber tubing
should be flushed with sample water to  remove air bubbles.   The  sample
bottle  and stopper should then be  rinsed thoroughly with sample water.
After rinsing,  the tubing  should be  inserted to the bottom of the sampling
bottle.   The bottle should be filled slowly until  at least half full, and
then filled rapidly thereafter.  At least one full  bottle volume of  sample
should  overflow the bottle before the tubing is removed.  After the tubing
is slowly removed,  the stopper should  be carefully put in place  with a
twisting  motion  while water is displaced from the  bottle.   Once stoppered,
the sample should be checked for air  bubbles.   If bubbles are present,
the sample should be discarded and a new sample collected.

     Processing—Because no reagents are used to preserve the oxygen samples,
analyses should be conducted Immediately after collection.   If a delay
occurs, it should be noted  on the log sheet.

Laboratory Procedures—

     Analytical Procedures—Detailed analytical  procedures  should be provided
by the manufacturer of the dissolved  oxygen  meter.  General  procedures
are listed in U.S. EPA Method 360.1 and APHA Method  421F.
                                    145

-------
                                                           Receiving Water
                                                   Dissolved Oxygen (Probe)
     Several  precautions should  be taken when  making measurements with
a membrane electrode.   First, constant  turbulence  should  be provided  by
a  stirrer to ensure  precise measurements.   Second,  adequate time should
be allowed for the instrument to warm  up before measurements are started
and,  when individual  samples are analyzed, for the probe to stabilize  to
sample temperature and  dissolved oxygen.   Third,  reactive  gases, such  as
chlorine and hydrogen sulfide, pass through the -"lembrane probes and may
interfere with  the analysis or desensitize  the probe.   Finally, broad variations
in th  kinds and concentrations of salts  in samples can  influence the  partial
press re of oxygen in  samples and thereby affect measurement accurscy.

     Calibration and Preventive Maintenance—Calibration procedures should
follow the instructions given by the manufacturer  of the  dissolved  oxygen
meter.   The meter generally can  be calibrated using one of three methods:
Winkler titrati'on, saturated water, or  air.   The air  method is simplest
and quite reliable.   Overall error  is diminished when the probe and instrument
are calibrated under conditions of temperature and dissolved oxygen  that
match those of the samples.  Calibration  can be disturbed by physical  shock,
touching the membrane,  or desiccation of  the electrolyte.

     Preventive maintenance procedures  should  follow the manufacturer1 s
reconnendations.  The oxygen probe should always be  stored in a humid environ-
ment to prevent drying  out and the  need to frequently replace membranes.

     Quality  Control  Checks—The  instrument should  be calibrated at the
beginning of each series  of measurements  and after each group of 10 successive
samples.   Duplicate measurements should be made on  at least  10 percent
of the total number of  samples.

     Corrective Action--If the dissolved oxygen meter does not appear  to
be operating correctly,  consult the  manufacturer's  troubleshooting guidelines
for remedial actions.
                                    146

-------
                                                            Receiving  Water
                                                   Dissolved Oxygen (Probe)
     Data Quality and Reporting—Repeatability of dissolved oxygen measurements
using a membrane electrode should be 0.1 mg/L and accuracy should be +1 percent
(J.S. EPA 1979b).  Sensitivity of  the  electronic  readout meter for the
output from the dissolved  oxygen probes should normally be 0.05 mg/L (U.S. EPA
1979b).   Dissolved oxygen  concentrations should  be reported  in mg/L to
tne nearest 0.1  unit.   Results should be reported for all  determinations,
including QA replicates.  Any factors that may have influenced sample qual ity
snould also be reported.
                                      147

-------
                                                           Receiving Water
                                                                       BOD
Biochemical Oxygen  Demand (BOD)

Field Procedures--

     Conection--BOD  samples can be collected in glass  or  plastic containers.
Sample containers and caps should be  rinsed thoroughly with sample  water
before sample collection.

     3rocessing--BOD  samples should be analyzed immediately  after collection.
If  a  delay occurs,  samples should  be  held at 40  c  to  min  lize reduction
of  BOO.  Samples should  not be stored  for more than  48 h and the  length
of  storage should be  recorded on the  log sheet.  Refrigerated  samples should
be  warmed to 2DO C prior  to  analysis.

Laboratory Procedures--

     Analytical  Procedures--BQD concentrations should be determined according
to  U.S. EPA Method 405.1  or  APHA Method 507.   Generally seawater  samples
require no dilution.   Samples having more oxygen-demanding  materials than
the amount of oxygen  in  air-saturated water should  be diluted to  balance
the oxygen demand and  supply.  If  samples are diluted, nutrient addition
(i.e., nitrogen, phosphorus, trace metals) and pH buffering  of the dilution
water are necessary to ensure that the sample  is suitable for bacterial
growth.  To prevent air  from infiltrating the  incubation bottles,  a water
seal  should  be  used.   When samples are incubating, all  light should be
excluded to prevent photosynthetic production of oxygen.  Samples containing
residual chlorine must  be dechlorinated  (e.g., using sodium thiosulfate).

      Calibration and  Preventive Maintenance—Dissolved oxygen  concentrations
should be  measured on all dilution water blanks and  seed controls  if appli-
cable.  Generally, seawater samples  require no dilution of seed.  Therefore
to  monitor  the  method,  a glucose-gultamic  acid  standard check  solution
should be  incubated with each  batch  of samples.  Dissolved oxygen measurements
                                     148

-------
                                                           Receiving  Water
                                                                       BOO
snould be calibrated according  to  accepted procedures (e.g.,  see descriptions
of the Winkler  and probe methods  in this document).

     Quality  Control Checks—The dilution water blank and the glucose-glutamic
acid standard provide quality control on  the  dilution water as well as
on  the  cleanliness of analytical  equipment  (e.g., incubation bottles).
Each sample should be analyzed  in  triplicate  to monitor precision.  Optimal
uatake  is at least 2 mg/L after the incubation, with a residual-oxygen
concentration of at least 1  mg/L  in  the sample.  Duplicate analyses should
be conducted  on a* '-east 10  percent  of the total number of samples.

     Corrective Action--APHA (1985)  should be  consulted  for methods of
correcting the  many kinds of interference that may  accompany  BOO analyses.

     If the dilution water blanks  exceed 0.2 mg/L,  cleanliness of containers
and water should be checked.  Containers may require 1+1 HC1  rinse  after
detergent washing to remove any residual organic  material.   Any containers
rinsed with acid should be thoroughly rinsed with  distilled water to prevent
acid carryover.

      If  a 2-percent  dilution of the glucose-glutamic acid standard  check
solution  is outside the  range  of 200 +. 37  mg/L,  BOD determinations made
with the seed and dilution water should be rejected.   Several methods used
to  determine the problem include  running a series  of dilution water blanks
using different water  sources  with and without  seed,  preparing a  fresh
 solution  of glucose-glutamic acid, changing the  seed,  or preparing  fresh
 reagents  for the dilution water.   The source of the problem  should be determined
before  additional  BOD  analyses are performed.

      Data Quality and Reporting — Precision data for spiked natural  waters
 indicate  that  standard deviations of  +0.7  and j+26 can be  achieved  for BOO
 concentrations of 2.1  and 175   mg/L, respectively (U.S.  EPA  1979b).   There
 is no acceptable  method  for determining  the accuracy  of the  BOD  test.
 BOD data  should be reported as mg/L to the nearest 0.1 unit.   Results  should

                                     149

-------
                                                            Receiving Water
                                                                       800
be  reported for all  determinations, including QA replicates,  dilution water

clanks, and glucost-gl utamic  acid  standards.   Any  factors  that  nay  have

influenced sample q-jal ity  should also be reported.
                                      150

-------
                                                           Receiving Water
                                                           Oil and Grease
Oil and Grease

Field Procedures--

     Conection--Samp1es  should be collected in glass bottles.  Bottles
should first be washed with a  warm  aqueous detergent mixture, and  then,
in  sequence, thoroughly rinsed with  hot tap water, rinsed at least twice
with distilled water, rinsed witn  1,1,2-trichloro-l,2,2-trifluoroethane
(i.e.,  Freon or  equivalent), ard  dried in a clean oven at ^i05o C for 30
min.  Bottle caps  should be lined  with  ' :E-coated cardboard inserts or aluminum
foil.   Plastic  containers are not acceptable.  Headspace should be left
in each sample container for  addition of  acid and mixing.

     Processing—Acidify  the  sample  in  the  collection bottle to a pH <2
using sulfuric or hydrochloric  acid.  Samples should  be stored  in the  dark
at  40  c.   Recommended maximum holding  time  is  28  days  (U.S. EPA 1984).
The length of storage should  be recorded on the log sheet.

Laboratory Procedures--

     Analytical Procedures--0il  and  grease  should be measured according
to U.S. EPA Methods 413.1  and 413.2  or  APHA Methods 503A and  5038.   Because
asphaltic materials are insoluble in Freon, recommended methods will give
low recoveries for samples  containing such material.   The gravimetric method
for oil  and grease measures relatively nonvolatile hydrocarbons, vegetable
oils,  animal fats, waxes, and  soaps.  The method is often used for wastewater
analyses because it requires minimal  instrumentation  and calibration.
Sulfur causes interference  for  the  gravimetric method because it  will  be
extracted and included as  oil and grease.  Light hydrocarbons that volatilize
at  temperatures  below 70<> C  (e.g.,  gasoline  through No.  2 fuel oil) are
lost during the solvent removal step.

     The  infrared oil and  grease procedure is more precise at lower concen-
trations  (1-10 mg/L) than  the gravimetric oil  and grease procedure.   The

                                    151

-------
                                                           Receiving Water
                                                            Oil  and Grease
infrared  method will give  higher results  than  the gravimetric method if
tne sample contains volatile components,  and  will give lower  results if
t-.e sample is high  in sulfur content.   A  limitation to the infrared determi-
nation is the method of standardization,  which  requires a prepared reference
oil.   The reference oil may not be comparable  to the type of oil  and grease
in the samples,  resulting in inaccurate results.

     Calibration and Preventive Maintenance—For gravimetric oil  and grease
analyses, check  accuracy of the analytical  balance periodically (minimum
of once per week rec  nmended)  using Class S weights.  A service contract
tnat includes schedulec preventive maintenance at least once  per  year is
recommended.  Scratched, chipped, or cracked  boiling flasks should be replaced.
For infrared oil and grease analyses,  follow  the manufacturer1 s  preventive
maintenance procedures  for the infrared  spectrophotometer.   Cells  used
for analysis should be  checked for  scratches  each time  they  are  used.
Scratched cells should  not be used.   For  infrared analyses,  calibration
should be performed as  specified in U.S.  EPA  Method 413.2, Section 6.4.

     Quality Control  Checks--Duplicate samples  should be collected and
performed at  a  minimum of  every 10  samples to establish  an  estimate of
precision.  Because samples should not  be split after collection, separate
grab samples should be  taken for analysis.

     Distilled  water spiked with a U.S.  EPA  performance sample or reference
oil should be extracted and  analyzed every  20 samples  to  monitor recovery.
Bottles  should be checked for cleanliness by analyzing  distilled  water
that has been acidified  in a  sample bottle.   A solvent  blank  should accompany
each  batch  of  samples.  A procedural  blank  should  be run with each  batch
of samples  to monitor reagent contamination or  procedural problems.

     Corrective Action—If oil  and grease concentrations  in  procedural
blanks are  greater than the  detection  limit, check the cleanliness of all
glassware.   Always use  separatory funnels with TFE stopcocks to avoid contam-
 ination from stopcock grease.   For  infrared  analyses, severe interferences

                                     152

-------
                                                           Receiving Water
                                                            Oil and Grease
will  result if Freon contacts  any material  containing plasticizers (e.g.,
plastic tubing).   For gravimetric analyses,  high results  will be  obtained
if any Freon or fumes remain in the flask  after  distillation.

     If difficulty occurs with  emulsions during  solvent  extraction, follow
procedures described in the methods.   If the emulsion  still  fails to dissipate
after addition of  salts, gently turn the separatory  funnel to  the horizontal
position and slowly rotate.   Be  careful  to keep the  stopper securely in
place.

     If precision or recovery  from spike results is p  or,  check adequacy
of extraction by increasing shaking time.

     Data  Quality and  Reporting--The definition  of oil and grease is based
on the procedures used.   Unless  identical procedures are used, oil  and
grease determinations are not intercomparable.  Therefore, the method  used
for analysis should always be specified.

     Objectives for precision and accuracy are indicated below:

     0     Reproducibility for the  gravimetric method using  distilled
           water spikes is ^18-percent coefficient of  variation  (U.S. EPA
           1983).   These  results probably  overestimate  the  precision
           possible with effluent samples.  No  precision data are reported
           by the U.S. EPA  for the infrared method.

      •    Inter laboratory recovery  results were  102+37  percent  at
           6.0 mg/L and 97+35 percent at 18.0 mg/L for  the gravimetric
           method  (U.S. EPA 1983).

      Measurements  should  be reported to a minimum of two significant  figures
 in mg/L.   Detection limits are in the  range of  5  mg/L  for gravimetric oil
and grease determinations  (U.S.  EPA  Method 413.1), and 0.2 mg/L for infrared
determinations (U.S.  EPA Method  413.2).  Results of all determinations

                                     153

-------
                                                            Receiving Water
                                                             Oil and Grease

should  be reported,  including QA  replicates, blanks,  spiked  samples, and
reference measurements.   Any factors that may have influenced  sample quality
should be reported.
                                     154

-------
                                                           Receiving Water
                                                        Nitrogen (Ammonia)
N'trogen (Ammonia)

Field Procedures--

     Conection--Ammonia samples can  be  collected in glass or  plastic bottles.
Each bottle and cap should be  rinsed thoroughly with  sample water  prior
to sample collection.

     Processing--Resuits of ammonia  analyses  are most reliable when they
.•-e made on fresh samples.  However,  if  analysis must ae  delayed, samples
can  be  stored for up to 28 days by acidification to pH<2 with sulfuric
acid  and refrigeration at 40 C.  The length  of  delay before  analysis should
be recorded on the log sheet.

Laboratory Procedures--

     Analytical  Procedures—Manual  distillation of effluent samples prior
to ammonia determinations is required by the  U.S. EPA, unless data on  file
demonstrate that distillation 1s not required.   Distillation  is recommended
due to the sensitivity of amnonia  procedures to color and possible interferences
in  the  effluent.  Following  distillation at pH 9.5, ammonia concentration
in saline water can be determined by  U.S.  EPA Methods  350.1, 350.2,  or
350.3 or APHA Methods 417A,  417B,  4170, or  417G.  The methods should  be
consulted to determine which is most  appropriate for available equipment,
expected concentrations, and expected levels  of  precision  and  accuracy.

      Calibration and Preventive Maintenance—Calibration procedures  should
follow  those  specified in  the method.  If samples  are being  distilled,
standards  should also be  distilled prior to  analysis  to check  for  ammonia
contamination or loss  during processing.  Concentrations of the calibration
standards  should bracket  the sample concentrations.   If  a  sample concentration
 is outside  the range of calibration, then an additional  calibration standard
should  be  analyzed  to check  if the  result is within the  linear range of
                                     155

-------
                                                          Receiving Water
                                                       Nitrogen  (Ammonia)
t.-e method.   Alternatively, the  sample  should be diluted  to within the
calibration range  and then reanalyzed.

     Quality  Control  Checks--Duplicate  analyses  should be conducted on
a minimum of 5 percent of the total number of samples,  with  an  additional
5 percent of the  samples spiked  and  analyzed for percent recovery.  A blank
snould be analyzed with  each batch  of samples.   A  U.S.  EPA performance
sample should be analyzed at least  once per quarter.

     Corrective Action—Contamination of ammoma  sample? can recur easily
due to  the volatile nature of ammonia.  To prevent possible  cro s-contamination,
reagents used  for other  analyses that contain ammonia (e.g., colorimetric
phenol) should  be  Isolated from samples and  standards used for  ammonia
determinations.  In addition, cleaning preparations  that contain significant
quantities of  ammonia  (e.g., Pinesol, wax  removers)  should not  be used
in  the laboratory area where ammonia  determinations  are  performed.

     Contaminated glassware should be  rinsed with 1+1 HC1 and then with
distilled water.  To  check for contamination,  blanks  should be analyzed
whenever  a new reagent  is  prepared.

     Data Quality and Reporting—Detection and  accurate  quantification
of  ammonia  in  receiving  water is routinely attainable, although method
detection limits can  vary widely because  of  methods or  instrumentation.
The analytical method should  be consulted  to determine expected  detection
 limits,  precision,  and  accuracy.  Data should be reported in mg/L as N
 to  a maximum of three significant  figures.  Results should be reported
 •or all  determinations, including QA replicates and spiked samples.  Any
 factors  that may have influenced sample quality  should  also be recorded.
                                     156

-------
                                                           Receiving Water
                                                 Nitrogen  (Total Kjeldah!)

Nitrogen (Total Kjeldahl)

Field Procedures--

     Collection--Kjeldah!  nitrogen samples can be  collected  in  glass or
plastic containers.  Caps  should  be  unlined, as  paper and/or glue may  interfere
with  the  analysis.  Each container  and cap  should  be rinsed thoroughly
with sample water prior to sample collection.

     Processing--I1  possible,  samples  should  be analyzed immediately after
collection.  If immediate analysis is not possible, samples  can be stored
up  to  28  days  by acidification  to pH<2  with  sulfuric  acid and refrigeration
at  40  C.   The  length of delay  before analysis  should be recorded  on the
log sheet.

Laboratory Procedures--

     Analytical  Procedures—Approved  test procedures  for the analysis of
total Kjeldahl nitrogen  in  saline water  include U.S.  EPA Methods 351.1
and 351.3 and APHA Methods  417B, 417D, 417E, 420A, or 420B.  The methods
should be  consulted  to determine which  is most appropriate for available
equipment, expected concentrations,  and  expected level's of precision and
accuracy.

      Calibration and Preventive Maintenance—Calibration procedures should
 follow those specified  in the method.   If  samples are  being digested and
distilled, standards should also be digested and distilled prior to analysis
 to check for ammonia contamination  or loss during distillation.   Concentrations
 in the calibration standards  should  bracket the  sample concentrations.
 If a sample concentration is outside of  the  range  of calibration,  then
 an additional  calibration standard should be analyzed  to check if the result
 is within the  linear range of the method.  Alternatively, the sample  should
 be diluted to  within the  calibration range and then reanalyzed.

                                     157

-------
                                                           Receiving Water
                                                 Nitrogen (Total  Kjeldahl)

     Quality  Control  Checks--Duplicate analyses  should be conducted on
a .-ninimum of 5 percent of the total  number  of samples, with  an  additional
5 percent of the samples spiked and analyzed  for percent recovery.  A blank
snould be analyzed  with  each batch of  samples.  A U.S. EPA  performance
sample should be analyzed at least once per  quarter.

     Corrective Action—Because ammonia is  a component of Kjeldahl  nitrogen,
precautions against  contamination that  were described for  anmonia  analyses
snould be followed.

     Data  Quality and Reporting—The  detection  and accurate quantification
of Kjeldahl nitrogen in receiving water is  routinely attainable,  although
method detection limits can vary widely because of methods or instrumentation.
The analytical method  should be consulted to determine  expected detection
limits, precision, and  accuracy.  Data should  be  reported  in mg/L as  N  to
a maximum of three significant  figures.  Results  should be reported for
all  determinations,  including QA replicates,  blanks, and  spiked samples.
Any factors that may have influenced sample quality  should also be recorded.
                                     158

-------
                                                           Receiving Water
                                            Nitrogen (Nitrate  and  Nitrite)
Nitrogen (Nitrate and Nitrite)

      Procedures--
     Conection—Nitrate-nitrite samples can be collected in  glass  or plastic
containers.  Prior to sample collection, each  container  and  cap should
be rinsed thoroughly with sample water.

     Pr oc ess ing--Ni irate- nitrite samples  should be  analyzed  immediately
after collection.   If a delay  occurs,  sa iples can be stored  for  up to 24  h
by  acidification to pH<2 with sulfuric acid  and held at  40  C.  Samples
must not be preserved using  mercuric chloride because the mercuric  ion
accelerates the degradation of the  cadmium-reduction column (APHA 1985).

Laboratory Procedures--

     Analytical  Procedures—Approved  test procedures for the  analysis of
nitrate-nitrite in  saline waters  include U.S.  EPA Methods 353.2  and 353.3
and APHA Methods 418C and 418F.  The methods should be consulted  to determine
which is most appropriate for  available  equipment,  expected concentrations,
and desired levels  of precision  and accuracy.

     Calibration and Preventive  Maintenance—Calibration procedures should
 follow those specified in the  method.   Efficiency of each  reduction column
should be checked by comparing  a nitrite  standard to a  reduced nitrate
standard  at the same concentration. This  efficiency check  should be made
at the beginning  and the end of each sample  run and at a minimum frequency
 of every  10 samples.  Reactivate the  copper-cadmium granules when  reduction
 falls below 75 percent.

      Concentrations of the calibration standards  should bracket  the sample
 concentrations.   If a sample concentration is outside  the  range of  calibration,
 then an  additional  calibration  standard  should be  analyzed to check  if

                                    159

-------
                                                           Receiving Water
                                             Nitrogen (Nitrate  and Nitrite)

the  result is within  the  linear range  of  tne method.  Alternatively, the
sample should be diluted to within the calibration range and reanalyzed.

     Quality  Control  Checks--Dup1icate analyses  should  be conducted on
a minimum of 5 percent  of the total  number of samples, with  an  additional
5 percent of the samples spiked and analyzed for percent recovery.  A blank
fiat has been run  through  the reduction column should be  analyzed  with
each  batch of samples or  every 20 samples.  A U.S. EPA performance sample
should be analyzed ?   least once per quarter.

     Corrective Action--Various components  of the effluent can  interfere
with the analysis.  The  method should be reviewed for ways to remove possible
interferences prior to analysis.  Possible intereferences include  suspended
solids, residual  chlorine, oil and grease, and high concentrations  of  iron,
copper, or other metals.

     The  area where nitrate-nitrite analyses are performed should be well
isolated from exposure  to nitric acid or nitric acid fumes.

     Data  Quality and Reporting—The detection and accurate quantification
of nitrate-nitrite in  receiving water  are routinely attainable,  although
method detection  limits  can vary because  of methods or instrumentation.
The analytical method  should be consulted to determine expected detection
limits, precision, and  accuracy.   Data should  be reported  in  mg/L as  N
to a maximum of three  significant figures.  Results of all  determinations
should be reported,  Including QA replicates, blanks, and spiked samples.
Any factors that may have influenced sample quality  should also be  reported.
                                     160

-------
                                                           Receiving Water
                                                        Phosphorus (Total)

Pnosphorus (Total)

Field Procedures--

     Co llect ion—Phosphorus  samples may  be  collected  in glass or plastic
containers.  Containers  should be  rinsed with  IN  HC1  followed by  several
rinses with distilled  water.  Detergents containing phosphate should never
be used on containers or labware that is to be used for  phosphate analysis.
Sample containers  and  lids should  be  rinsed thoroughly  rith sample water
before sample collection.

     Processing--Phosphorus  samples can  be  stored up to  28 days before
analysis  by acidification to pH<2  with sulfuric  acid  and  refrigeration
at 40 c.   Samples with low concentrations of phosphorus should not be stored
In plastic containers, as  phosphates may adsorb onto the'container walls.

Laboratory Procedures--

     Analytlcal Procedures—Approved test  procedures  for  the analysis of
total phosphorus in receiving water are U.S. EPA Methods  365.1, 365.2,
and  365.3 and  APHA Methods 424C,  424F,  and 424G.  The methods should be
consulted to determine which is most appropriate for available equipment,
expected  concentrations, and desired  levels of precision  and  accuracy.

     Calibration and Preventive Maintenance—Calibration procedures should
follow those specified in  the method.   Concentrations  of the calibration
standards should bracket the sample concentration.   If a  sample concentration
is outside the range of  calibration, then an additional calibration standard
should be analyzed to  check  if the result is within  the  linear range of
the method.  Alternatively, the  sample should be diluted  to within  the
calibration range and reanalyzed.

     Quality Control Checks—Duplicate analyses should be conducted on
a minimum of 5 percent of  the  total number of samples,  with  an additional
                                    161

-------
                                                           Receiving Water
                                                        Phosphorus (Total)

5  percent of the  sanoies spiked and analyzed  for percent recovery.  A blank
should be analyzed  with  each batch of  samples.   A U.S. EPA performance
sample should be analyzed at least once per quarter.

     Corrective Action — Because  phosphorus contamination can occur from
a  variety of sources,  it is recommended that  a clearly marked set of  lab-
ware  be  dedicated  to  only  phosphorus analysis.  This labware should never
be exposed to pnosphorus detergents or reagents containing phosphate.

     Various  components  of the  effluent can interfere witt the analysis.
The method should  be reviewed for  vays to remove  interferences or  adjust
for interferences  from  components that cannot  be removed.  Silica and  arsenic
are possible positive  Interferences, while hexavalent chromium and nitrite
can cause low recovery.

     For  highly colored  or turbid  samples, additional sample preparation
(e.g., further oxidation  or filtration) may be  required prior to  color
development.  In  any  case,  blanks should be prepared by adding all  the
reagents except the  coloring reagents to  the  sample.  Measure absorbance
in  the  sample blank at the wavelength used for the phosphorus determination
and subtract this absorbance value from the sample absorbance prior  to
calculation of phosphorus concentration.

     Data Quality and Reporting—Detection and accurate  quantification
of total  phosphorus  in  receiving water  is routinely attainable.   Actual
method detection  limits can  vary because of methods or instrumentation.
The analytical method  should be consulted  to  determine expected detection
limits,  precision, and  accuracy.  Data should  be reported in mg/L as P
to a maximum of three  significant figures.  Results of  all determinations
should be reported,  including QA replicates and spiked samples.  Any factors
that may have influenced sample quality  should also be reported.
                                     162

-------
                                                          Receiving  Water
                                         Total  and Fecal Coliform Bacteria
Total and Fecal Coliforro Bacteria
Field Procedures--

     Con ection--Samples  should be collected  in  clean,  sterile polypropylene
or glass containers.  The sample containers must be  resistant to sterilizing
conditions and to the solvent action  of water.  The  container lids must
not  produce bacteriostatic or nutritive compounds  upon  sterilization.
The  sample containers must seal  tightly.   Containers  with chips, cracks,
or etched marks should be discarded.

     Heat-resistant glass  or plastic sample  containers should be autoclaved
at 1210  c for 15 min.  Alternatively, dry glass containers can be sterilized
in  a hot-air oven at 1700 c for at least 2  h.  For plastic containers that
are  not  heat-resistant, ethylene  oxide gas  sterilization is acceptable
(Bordner et al.  1978).  Containers  sterilized by gas  should be  stored at
least 12 h before use to ensure all gas  has dissipated.

      If  the sample water  has  residual chlorine, sodium thiosulfate  should
be  added (to a concentration of 0.008 percent)  to neutralize the  chlorine
and  thereby prevent continued bactericidal  action after sample collection.
 In  this  manner, the  true  microbial  content  of the water at the time of
sampling  can  be  estimated more  accurately.   If  sodium thiosulfate must
be  added to a  sample, it  should  be added  to  the sample container prior
to  sterilization.

      If the sample water  contains heavy  metals in concentrations  exceeding
 0.01 mg/L, a chelating agent  should be added to the sample  container  to
 reduce  metal  toxicity.   This  is  particularly important if samples are  not
 analyzed within 4 h after  collection.  APHA  (1985)  recommends  using  the
 disodium  salt of ethjlenediaminetetracetic acid (EOTA), adjusted to pH
 6.5, and added  to the sample container before  sterilization.   For  a  120-mL
 container, addition of 0.3  ml  of a 15-percent EDTA  solution is considered
 adequate (APHA 1985).

                                    163

-------
                                                           Receiving Water
                                         Total and Fecal  Coliform  Bacteria
     It is  critical that samples are not  contaminated during the collection
process.  To avoid contamination,  sterilized containers  should be kept
sealed  until they are used,  containers  should be filled without rinsing,
and container lids  should be replaced  immediately after  the  samples have
been  collected.  When removed  from containers,  lids should be held face
dcwn in one hand and not set down on any surface.  Adequate headspace (at
least  2.5 cm)  should be left in each sample  container to facil itate  mixing
prior to analysis.

     Processing—Samples should be analysed  as  soon as possible after col-
lection.   If a delay occurs, samples should be held at 40 C for a maximum
of 6 h.  The length of delay should be noted on the  log sheet.

Laboratory Procedures--

     Analytical Methods—Details of the membrane filter (MF) method  and
the most  probable number  (MPN) method are  presented  in  Part III of Bordner
et  al. (1978)  and in  Parts 908  and  909 of  APHA (1985).  Although  the  MF
method  is more  precise  than  the MPN  method,   it  is  also  more sensitive  to
 interference  from turbidity in  samples.  Because  the MF technique usually
yields  low  and  variable recovery  from chlorinated waters,  the MPN  technique
should  be used  when samples  contain  chlorine  residual.

      Calibration  and Preventive Maintenance—This information  is reviewed
extensively  in  Part IV of Bordner et al. (1978) and  in Part 902 of APHA
 (1985).

      Quality Control Checks--Quality control checks  for total and fecal
 coliform  bacteria  analyses  are listed  in  detail  in Part IV of  Bordner et al.
 (1978) and  in  Part 902 of APHA (1985).  The list includes:

      •    Sterility checks  on media, dilution and rinse water, glassware,
           and membrane filters

                                      164

-------
                                                          Receiving Water
                                         Total and Fecal Coliform Bacteria
     •     Duplicate analyses on 10 percent of samples and  on at least
          one sample per test run

     •     Colony verifications on a monthly basis.

     Corrective Action—Procedures detailed in  the  relevant  sections of
Bordne:  et al.  (1978) and APHA  (1985) should be followed.

     Data Qua! ity and  eport ing—Table 909:11 of APHA (1985) presents 95-percent
confidence limits for - ne MF  method  for coliform colonies of  1,  2, 3, 4,
5 and 10.   The precision of  the  MPN method increases with increasing number
of replicates.  With five tubes, each with  1 ml of  sample, a  completely
negative, result  is expected  less  than 1 percent of the time  (APHA 1985).
Confidence limits (95 percent)  for various MPN counts  are  given in Tables
908:111, 908:IV, and 908:V of APHA (1985).

     Using  the MF  method, data should  be reported  as densities  of coliforms
per 100 mL.  Using  the MPN method,  data should be reported  as  MPN values
per  100 ml.  Results of all  determinations should  be  reported, including
QA replicates,  sterility checks, and  colony verifications.   Any  factors
that may have influenced sample quality should also be  reported.
                                     165

-------
                                                           Receiving Water
                                                     Enterococcus  Bacteria
Enterococcus Bacteria

Field Procedures--

     Col lection—Samples snould be collected  in clean, sterile polypropylene
or glass containers.  The sample containers must be resistant to sterilizing
conditions and to the solvent action of water.   The container lids must
not  produce bact :riostatic  or nutritive compounds upon  sterilization.
The  sample containers must  seal  tightly.   Containers with chips,  cracks,
or etched marks should be discarded.

     Heat-resistant glass or plastic  sample  containers should be autoclaved
at 121Q C for 15 min.  Alternatively,  dry  glass containers can be sterilized
in a hot-air oven at 1700 C for at  least 2  h.  For plastic containers that
are not heat-resistant,  ethylene oxide  gas sterilization  is acceptable
(Bordner  et al. 1978).  Containers sterilized by gas should be stored at
least 12 h before use to ensure all gas  has dissipated.

     If the  sample water has  residual chlorine, sodium thiosulfate should
be added to neutralize the chlorine and  thereby prevent continued bactericidal
action  after sample collection.  In  this  manner, the true microbial  content
of the water at the time of sampling can be estimated more accurately.
If sodium thiosulfate must  be added to a  sample, it should be  added to
the  sample container prior to sterilization so that the final concentration
in the sample will  be 100 mg/L.  For  a 120-mL container, 0.1 nt of a 10-percent
solution of sodiun  thiosulfate will neutralize a sample containing  as  much
as 15 mg/L of residual chlorine (APHA  1985).

     If the  sample water contains heavy  metals in concentrations exceeding
0.01 mg/L, a chelating agent should be added to the sample container to
reduce  metal toxicity.  This is particularly important if samples are not
analyzed within 4 h after collection.  APHA (1985) recommends using the
disodium  salt of  ethylenediaminetetracetic acid (EOTA),  adjusted  to pH
                                     166

-------
                                                           Receiving Water
                                                     Enterococcus Bacteria
6.5,  and  added to the sample  container before sterilization.   For a 120-mL
ccnta-'ner, addition of 0.3  ml of a  15-percent EDTA solution is  considered
acequcte (APHA 1985).

      It is  critical that samples are not contaminated during the collection
process.  To avoid contamination, sterilized  containers  should  be  kept
sealed  until  they are used,  containers should be filled  without rinsing,
and container  lids  should be replaced immediately after the  samples  have
been  collected.  When removed from containers,  lids should be held face
down  in one hand and not set down on  any s rface.   Adequate headspace (at
least 2.5 cm) should be left  in each sample container to facilitate mixing
prior to analysis.

      Process ing--Samples  should be  analyzed as soon as possible after col-
lection.   If  a delay occurs, samples should be held at 40 C  for a maximum
of 6  h.  The length of delay should be noted on the log sheet.

Laboratory Procedures--

      Analytical  Methods—Methods for  analyzing enterococcus bacteria are
currently being finalized by U.S. EPA.

      Calibration and  Preventive  Maintenance—This information is  reviewed
extensively in Part IV of  Bordner  et al. (1978)  and  in Part 902  of  APHA
(1985).

      Quality  Control  Checks—Quality  control checks for these analyses
are  listed  in detail  in Part IV of  Bordner  et  al. (1978)  and in  Part 902
of APHA (1985).  The list includes:

      •    Sterility checks on media,  ailution and  rinse water, glassware,
          and membrane filters
                                     167

-------
                                                            Receiving Water
                                                      Enterococcus Bacteria
     •    Duplicate  analyses on 10 percent of samples and  on  at  least
          one sample  per  test run

     •    Colony verifications on a monthly basis.

     Corrective Action--Procedures detailed in the  relevant sections of
Bordner et al.  (1973)  and APHA (1985) should be followed.

     Data Quality and Reporting — Data should be reported  according  to the
specifications  v. -rently  being finalized by U.S. EPA.
                                      163

-------
                                                           Receiving Water
                                                             Chlorophyll  ^
Chlorophyll a

Field Procedures—

     Col lection—Chlorophyll  ^ samples  can be collected in glass or plastic
containers.  Because acids  will  decompose chlorophyll a_ to  phaeophyton,
it  is  critical  that the water  sampler and  sample containers remain free
o* acids (including  acids frcm  fingerprints).   In addition, sample containers
should be rinsed  three times before sample collection.

     Processing—Chlorophyll  ^ samples  should oe  filtered immfcdiately after
collection.  Two  or three drops  of magnesium chloride suspension  should
be added to the sample before  filtration to prevent the sample from becoming
acidic.  Filters  can be stored for a few weeks by  holding  them in the  dark
in  a desiccator  at -200 c (Strickland  and Parsons 1972).  However, storage
usually leads to  low results and  makes the extraction of  chlorophyll  more
difficult.  It is therefore recommended  that filters be extracted imnediately
after filtration.   If filters  are stored, the  length of delay until analysis
should be recorded  on the log  sheet.

Laboratory Procedures--

     Analytical Procedures—Chlorphyll £ determinations using the fluorometric
method are described in detail in Section IV.3.IV  of Strickland and Parsons
(1972)  and as Method 1002G2 of APHA  (1985).  All  work with  chlorophyll
extracts should be  conducted in subdued  light  to avoid degradation.   Opaque
containers or containers  wrapped in  aluminum should be  used to protect
samples from light.  Glass fiber  filters are  recommended  because they are
inexpensive and  result  in practically no blank (Strickland and Parsons
1972).  To improve  extraction  efficiency,  a  cell-grinding steo  should be
included  prior to  extraction.  A TFE/glass grinder  should be used for glass
fiber filters.
                                     169

-------
                                                          Receiving  Water
                                                            Chlorophyll £
     Calibration  and  Preventive Maintenance—The fluorometer  should  be
calibrated spectrophotometrically using samples  from the same standard
chlorophyll  solution.  This  solution should have a  known  concentration
of chlorophyll  a^ that was extracted  from marine phytoplankton.  Strickland
and  Parsons  (1972) recommend  that a mixed culture  of equal  amounts  (by
pigment) of Skeletonema costatum, Coccolithus  huxleyi i, and Peridinium
trochoidium be  used as the sources  of  chlorophyll.  If  natural phytoplankton
populations are used, phaeo-pigments may be present.

     A  series  of  dilutions of the standard chlorophyll  solution  shu  Id
be made so that concentrations  of 2, 6, 20, and 60 ug/'L are achieved.
Readings of each dilution  should be made  at sensitivity settings of  Ix,
3x,  lOx, and 30x..   This will  allow derivation of  calibration factors  to
convert fluorometric readings in each sensitivity level  to chlorophyll ^
concentrations  (for details, see APHA  1985).

     Quality Control Checks—To correct  for  scatter, the fluorometer should
be zeroed against  a cuvette of 90-percent acetone for each  level of sensi-
tivity.  Duplicate analyses  should  be conducted on at least 10 percent
of the  total  number of samples.

     Corrective Action—If the fluorometer does not appear to  be functioning
properly, the manufacturer1 s troubleshooting  guide should be consulted.
Scratched or etched cuvettes should be replaced.

     Data Quality and  Reporting—It is recommended  that chlorophyll £ concen-
trations be determined  using the fluorometric  method rather than  the spectro-
photometr ic method, because the former technique  is more sensitive, requires
less sample, and can be used  for in vivo measurements (APHA 1985).  Precision
of the  recommended method varies as a  function of  the amount  of pigment
being measured.   For chlorophyll £ concentrations  exceeding 0.5  mg/m3,
a  precision of +8 percent  is possible.   The sensitivity  of  detection  has
been estimated as 0.01 mg/m3 for a  2-L sample (Strickland and Parsons  1972).
Chlorophyll ^ concentrations should  be  reported as mg/m3 to the nearest

                                     170

-------
                                                          Receiving Water
                                                            Chlorophyll a

0.01  unit.   Results of all determinations should  be  reported, including

QA replicates.  Any factors that  may have influenced sample quality  should

also be reported.
                                       171

-------
                                                           Receiving  Water
                                                             Phytoplankton
Phytoplankton

Field Procedures--

     Pre-Collection  Preparation--A solution of  100-percent buffered  formalin
should be prepared by  adding 2 g of  sodium borate per 98  ml  of undiluted
formalin.   Sample jars  should have a  capacity of approximately 1  L and
should have lined plastic or polyethylene lids.   Each jar  should be marked
at the point where  it  is 90 percent full.

     Coliection--When the sampler  is  retrieved,  it  should be shaken  to
disperse the organisms.   (Note:  swirling is not effective for  dispersing
plankton.) Each sample should then be drained  into a marked 32-oz jar  until
the water level reaches the 90-percent full mark.

     Process ing--Because preservation renders many  phytoplankton  forms
taxonomically Intractable and  indistinguishable from detritus, it is highly
desirable to analyze  live material (Stofan and Grant 1978).  Unfortunately,
this is not practical  for most routine monitoring programs.   Hence, procedural
recommendations given  herein address preserved  samples.

     It  is recommended that formalin be used to fix the phytoplankton samples.
However, other fixatives  are available  if formalin proves unsatisfactory
(see  Stofan and Grant 1978).  Once a  fixative is selected, it should  be
used exclusively so  as not to  bias the data.

     After each sample has been  collected,  the  jar  should be  filled  to
the top  with 100-percent  buffered  formalin and the cap  should  be  screwed
on  tightly.  The sample jar  should then be  inverted several times to mix
the contents.

     Samples should be stored  upright in an opaque  box to  minimize exposure
to sunlight and in a cool area  to minimize exposure  to  high temperatures.
Samples should  also  be  stored in  a  stable part of the ship  to minimize

                                    172

-------
                                                           Receiving Water
                                                            Phytoplankton
agitation.   If special  racks are not available to keep the  shipping boxes
from moving, they  should  be  lashed down.

Laboratory Procedures--

     Analytical Procedures—As with most biological field samples, handling
is detrimental to specimen quality,  and hence,  data quality.   Handling
should therefore be kept  to  a minimum.

     Several methods  of  : >ecimen identification and  enumeration may be
used (Stofan and Grant 19/8).   The  most common  is  the Utermohl method,
by which  phytoplankters are identified using an inverted  microscope and
a fixed, settled sample.   Contents of the sample jar are first resuspended
by gently shaking in an up-and-down  motion.  An aliquot Is then poured
into  a graduated vertical counting  chamber and allowed to settle.  Phytoplank-
ters are then identified  as  they  rest on the bottom of the chamber.  Scanning
electron microscopy and  Nomarski  illumination  for  light microscopy can
be used to aid  taxonomic  identifications.

     If possible, phytoplankters should be identified and  enumerated from
an unconcentrated sample.  This 1s highly desirable because  sample  handling
is minimized.  However,  in  nutrient-poor waters the phytoplankton community
may be sparsely distributed, necessitating sample  concentration.   Several
concentrating procedures are commonly used  (see Stofan  and Grant 1978).
If sample concentration  is necessary, tests  should be conducted  to see
which  method Is the least harmful  to the specimens.  The  selected method
should then be evaluated for efficiency by comparing it with  replicate
aliquots of unconcentrated  samples.  After being accepted, the method should
be the only one used for  concentrating  samples.  Data generated from concen-
trated samples should always be  evaluated in light of the results of the
efficiency tests recommended above.

     All phytoplankers should  be  identified to  the species level and  enumerated,
and  the data should be recorded on  a  log  sheet.  After  the  identifications
                                    173

-------
                                                            Receiving Water
                                                             Phytoplankton
h = ve  been completed  and the data  have been recorded,  each sample aliquot
trat was examined should be archived.   Each  aliquot should be  stored in
a separate vial  that is labeled  internally with the following  information:
survey area; station number; replicate number (if any);  latitude and  longitude,
Loran C, or other positioning  information; date of collection; and collection
depth.  Each vial should be sealed  tightly with  a  polyseal cap,  and all
vials from the  same  sample should  be  stored in a  common  container filled
^th preservative.  To prevent evaporation of the fixative, vial and container
lids  can  be sealed with plastic tape.  Although preserving and archiving
the sample aliquots is recommendeo for reference and  verification  purposes,
it  should be remembered that many  of  the organisms will  deteriorate over
time because of the caustic properties of the fixative.

     Calibration and  Preventive Maintenance—The inverted microscope should
be  serviced annually unless the manufacturer recommends otherwise.

      Taxonomic identifications should be consistent within  the  given  laboratory,
and with the identifications of other workers.   To that end,  at  least  three
individuals of  each  taxon should  be  sent for verification to  recognized
experts in museums and/or academic institutions.  The verified  specimens
snould then be  photographed, and  the  photographs  should  be placed in the
permanent reference museum.  Continued  collection  of a verified  species
does  not require additional expert verification.  Participation of the
laboratory staff in a regional  taxonomic standardization program (if  available)
is  required to  ensure regional consistency and accuracy  of  identifications.
The photographic reference collection should  be  started and completed as
early as  possible during the  project.

      Photographs of phytoplankton  taxa are recommended because of  the diffi-
culties  involved in preserving  phytoplankton  specimens.   The  photographs
should  be recorded on 35-mm color  slide  film using a high-quality  photomicro-
scopy apparatus.  Each  slide  should be coded,  and  an  accessory logbook
of  information  should be maintained.  The logbook  should  contain the  following
 Information:  survey name; station number; replicate number (if any);  latitude

                                     174

-------
                                                           Receiving Water
                                                            Phytoplankton
and longitude, Loran C, or other positioning information;  date of collection;
kind of sampler; and name of collector.  In addition, the name  of the  taxonomic
expert who verified the identification, institutional affiliation, and
tne date  of verification should be recorded.

     A computer  listing  of  each  species'  name, the person  who  verified
the identification, date of  verification, location of the photograph in
tne collection, status of  the photograph if it has been loaned to outside
exoerts,  and references to pertinent literature should be maintained by
tne laboratory performing  identification?  for the monitoring  program.
This listing should be available near the col ection.  Reference photographs
are invaluable,  and should be retained at the  location where the identifications
were performed,  in the offices of the funding  agencies,  or at  a  museum
with  long-term storage capabilities.   In no instance should this portion
of the collection be destroyed.  A  single person should be  identified as
the curator of the collection and be  responsible  for its integrity.

     Quality  Control  Checks--"New" taxonomists should have all of their
identified organisms verified within  the laboratory, until  their  accuracy
reaches  95 percent.  Thereafter,  all  identifications in one of every 10
samples should be checked for accuracy.

     The most  recent taxonomic  literature should be used whenever  possible.
Citations for all taxonomic references should be  appended to the data report
for each survey.

     Corrective Action—  If a  sample  is  found  in which one or more  species
are misidentified,  all previous  unchecked  samples  should  be  examined for
those specific errors.  Errors should be rectified when located.

     Data  Quality  and Reporting  Requirements--Al1  phytoplankton should
be  identified to the lowest possible taxon,  preferably to the species  level.
In  cases where  the identity of  a species  is  uncertain, a species number
may be used  (e.g.,  Chaetoceros sp. 1).   Data   for each sample  should be

                                    175

-------
                                                            Receiving Water
                                                              Phytoplankton

resorted  as number  of individuals  per liter  for  each species.  Results

of all determinations should  be reported, including taxonomic verifications.
Any factors that may  have influenced  sample quality should  also be reported.
                                     176

-------
SEDIMENT/INFAUNAL ANALYSES

     QA/QC procedures are  presented  in  this  section for the following  11
sediment/infauna variables:

     •    Grain size
     t    Total solids/water content
     •    Total volatile sol ids
     •    Total organic carbon
     •    Biochemical  oxygen demand
     •    Chemical oxygen demand
     •    Oil and grease
     t    Sulfides
     0    Priority pollutant metals
     •    Priority pollutant organic  compounds
     •    Infauna.

     Samples to be analyzed for these variables generally will be collected
 using  a  bottom  grab.  Operation of a  bottom grab is  discussed earlier  in
 the general methods section.
                                       177

-------
                                                                  Sediment
                                                                Grain  Size
Grain Size

Field Procedures--

     Col lection--Grain size  samples can be  collected in glass or plastic
containers.

     Processing — Grain size  samples should  be  stored in a refrigerator
&~  40  c to minimize  the  effects of bacterial growth.   They  may be  held
for up  to 6 mo in this manner.  Samples  must not be  frozen or dried p -ior
to  analysis, as either  of these processes  may change  the  particle-size
distribution.

Laboratory Procedures--

     Analytical Procedures—Grain  size should  be measured according  to
the sieving and pipet  procedures described in  Plumb  (1981).  A wet sample
is  first sorted into  coarse and fine  fractions by wet  sieving through a
63-urn (i.e., 4-phi)  sieve.  Particles  having a diameter greater than  63
urn  (i.e., -sand and  gravel)  are sorted by dry sieving  through a graded series
of  screens.  Particles  having a diameter less than  63  urn (i.e., silt and
clay)  are sorted by  their different  settling velocities during the  pipet
analysis.

     An aliquot of the wet-sediment sample  should be analyzed for  total
solids  to estimate the dry weight of the aliquot used  for  grain  size analysis.
This estimated weight  can  be compared with the summed weight of all fractional
determinations to evaluate the efficiency of the method.

     It is critical  that each  sample be thoroughly homogenized in the laboratory
before  aliquots are removed.  Laboratory homogenization  should be conducted
even if samples were homogenized  in the field.
                                     178

-------
                                                                  Sediment
                                                                Grain Size
     If organic matter is oxidized prior to size analysis, the true particle
size will be determined.  If organic matter is not oxidized prior to analysis,
tne apparent particle size will be determined.  Because results from these
two techniques can differ, one method should be  selected and used for  all
grain-size determinations.

     At a minimum, it is recommended that each sample be analyzed for  percent
gravel,  sand,  silt, and clay.  However, if the fine-grained fraction comprises
less  th*n 1  percent of the total  sample weight, the pipet analysis is not
recommer ed and the weight of the  combined  silt-clay  fractions  should be
reported.  In  very sandy sediments,  the sand  fraction may be subdivided
further.  If this is desired, it is  recommended that the  fraction  be  sieved
at  1-phi increments to yield five  subfractions.

     Calibration and Preventive Maintenance—The analytical  balance, drying
oven, sieve shaker,  and  temperature bath  should be  inspected  each  time
they  are used,  and calibrated weekly at a  minimum.   The  manufacturer's
instructions should be consulted for calibration of the  analytical balance
and for preventive maintenance  procedures for  the analytical  balance, drying
oven,  sieve shaker, and temperature  bath.  It  is recommended that a  calibrated
dial  thermometer be used in the drying  oven  to determine  average temperature.

      Quality Control  Checks—Several procedures  are critical to the collection
of  high quality particle size data.   Most  important  to  the dry sieve analysis
 is  that the screens are  clean before conducting the  analysis, and  that
 all of the sample  is  retrieved from them.   To clean a screen,  it  should
be  inverted and tapped  on a  table, while  making sure that  the rim  hits
 the table  evenly.   Further cleaning of brass screens may be  performed by
gentle scrubbing  with a  stiff bristle  nylon brush.   Stainless steel  screens
may be cleaned with a nylon or brass brush.

      The most  critical aspect of  the pipet analysis  is  knowledge of the
 temperature of the  silt-clay suspension.   An increase of only 1° C  will
 increase the settling velocity of a particle 50 urn in diameter by 2.3 percent.
                                    179

-------
                                                                 Sed imen t
                                                               Grain Size

It is generally recommended that the pipet  analysis be conducted at a constant
temperature of 200 c.  However,  Plumb (1981)  provides a table to correct
for settling velocities  at other  temperatures.  Thorough mixing  of  the
silt-clay suspension at  the  beginning of  the analysis is  also critical.
A perforated, plexiglass disc plunger is very effective for this purpose.

     Replicate sieve and pipet analyses  should  be  conducted on at least
10 percent of the  total  number of  samples analyzed.   If  >20 samples  are
sorted, it is recommended that at least one tripiic te en'lysis be performed.

     Corrective Action--An analysis should be repeated if weighing errors
exceed 5 percent of the original sample weight.   If  the mass of  sediment
used  for pipet analysis  exceeds 25 g (0.9  oz), a subsample should be used
as described by Plumb  (1981).  Silt-clay samples in excess of 25 g  (0.9 oz)
may  give erroneous results because  of electrostatic  interactions between
the particles.   Silt-clay samples less than 5 g yield  a  large  experimental
error  in weighing  relative to the total  sample weight.  Other reasons for
erroneous results include hindered settling and flocculation.

     Data  Quality and Reporting — Ideally,  the summed weight of all sample
fractions should differ  from  the original sample weight  by less than  1  percent
(Plumb 1981).  Weights  of the six gravel/sand fractions, the  total  silt
fraction, and the total  clay fraction should be  reported to the nearest
0.0001 g.  See Shepard  (1963) for  size  ranges of  the  textural  classes.
Results of  all determinations should be reported, including QA replicates.
Any  factors that may have  influenced sample quality should  also be reported.
                                     180

-------
                                                                  Sediment
                                                Total  Solids/Water Content
Total Solids/Water Content

Field Procedures—

     Collection--Samples can be collected  in glass or plastic containers.

     Processing — Samples should be  stored frozen  and  can be held for  up
to 6 mo *:n that condition.
Laboratory Procedures—

     Analytical  Procedures—Total  solids should  be determined using  the
procedure described in  Plumb (1981).   The general procedure is to weigh
a  wet sample, dry  it at  103-1050 C,  reweigh  the  sample, and determine  the
percentage of wet weight accounted for  by the dry weight.  Unrepresentative
material should  be removed from the  sample  prior  to  analysis and noted
on the  laboratory log sheet.

      Because total  solids  content is operationally defined by the drying
temperature,  it  is essential that  this temperature be held constant at
the specified value.   It is also  critical  that  each sample be thoroughly
homogenized  in the laboratory before  a subsample  is taken for analysis.
Laboratory homogenization  should be conducted  even  if  samples were homogenized
 w the field.  Ignition  of  the evaporating dish  prior to analysis  is critical
to ensuring  that the container is free  from contaminants.

      Calibration and Preventive Maintenance—The analytical  balance, drying
 over, and muffle furnace should  be inspected  and calibrated  weekly, at
 a minimum.

      Quality Control  Checks—Duplicate  analysis should  be conducted on
 at least 10 percent at the total  number of samples.

      Corrective  Action—None applicable.

                                      181

-------
                                                                 Sedimen t
                                                Total  Solids/Water Content
     Data Quality  and tteporting--Total  solids content or, alternatively,
water content should be reported  as  a  percentage  of  the wet weight of the
sediment  sample to the nearest  O.i  unit.  Results  of all determinations
snould be reported,  including QA replicates.   Any  factors that may have
influenced sample quality should  also  be reported.
                                    182

-------
                                                                 Sediment
                                                     Total  Volatile Solids
Total  Volatile Solids (TVS)

Field Procedures--

     Collection—Samples can be collected in glass or plastic  containers.

     Processing—Samples  can  be stored  frozen and  can  be  held  for  up to
6 mo in that condition.

Laboratory Procedures--

     Analytical Procedures—Methods  for volatile solids analysis  are  given
by Plumb (1981).  The basic procedure  is to dry a sediment sample at 103-1050 C,
weigh  it, and then combust it  at  5500 C.  The difference in mass  between
the combusted sample and  the desiccated  sample estimates the mass of volatile
organic matter,  unrepresentative material  should be removed from the sample
prior to analysis and noted on  the laboratory log sheet.

     Because total  volatile  solids  is operationally defined by the drying
and ignition temperature, it is essential that  these temperatures be held
constant at their specified  levels.   It  is also critical that each sample
be thoroughly homogenized  in the laboratory before a subsample  is  taken
for analysis.  Laboratory homogenization should  be conducted even if samples
were homogenized in the  field.   Evaporating  dishes  (or crucibles) must
be  ignited  at 550<> C before being  used for total volatile solids  analysis.
This step ensures that the dishes  are free  from volatile  contaminants.
Dried  and  combusted samples should  be cooled in a desiccator and held there
 until  they  are weighed.   If a  desiccator  is  not used,  the  sediment will
accumulate  ambient moisture  and  the sample  weight will be overestimated.
A color-indicating desiccant  is  recommended  so that  spent  desiccant can
be  detected easily.  Also, the seal on  the  desiccator should be checked
 periodically and,  if necessary,  the  ground  glass rims  should be  greased
or  the  "0"  rings should be replaced.
                                     183

-------
                                                                 Sediment
                                                     Total Volatile Solids

     Calibration and Preventive Maintenance--The analytical balance, drying
oven, and muffle furnace  should be inspected  and  calibrated weekly,  at
a minimum.   The muffle furnace thermocouples and  thermometers should  be
inspected before each use and calibrated as needed.

     Quality Control  Checks—Duplicate  analyses should  be  conducted  on
at least 10  percent of the total number of samples.

     Corrective  Action—None appl -cable.

     Data Quality  and  Reporting—Total  volatile solids  content should  oe
reported as  a percentage  of the dry weight  of the  uncombusted sediment
sample  to the  nearest 0.1  percent.  Results  should  be  reported for all
determinations, Including QA replicates.  Any factors that may have influenced
sample quality should also be reported.
                                     184

-------
                                                                Sediment
                                                     Total  Organic Carbon

Total  Organic Carbon (TOG)

Field Procedures--

     Collection—Samples  can be collected in  glass or plastic containers.

     Processing—Samples  should  be stored  frozen  and can be held  for  up
to 6 mo in that condition.  Excessive  temperatures  should  not be used  to
tnaw samples.

Laboratory Procedures--

     Analytical  Procedures—The  method  for total  organic carbon analysis
is given by Plumb  (1981).  The recommended procedure  is  to first acidify
the  samples to  remove inorganic carbon, and then convert all organic  carbon
compounds to carbon dioxide by catalytic combustion.

      Because  inorganic  carbon  (e.g.,  carbonates,  bicarbonates, free C02)
will  interfere with total organic  carbon determinations,  samples must  be
treated to remove inorganic carbon  before being  analyzed.   Samples may
be combusted in one of several  different models  of induction furnace.
All  combustions should be performed  in  a single furnace (even if two furnaces
 are  the same model),  as difrsrent furnaces may  give slightly  different
 results.   It is also critical that  each sample be thoroughly homogenized
 in the laboratory before a subsample  is  taken  for  analysis.  Laboratory
 homogenization  should  be conducted  even  1f samples were homogenized in
 the field.  Dried samples  should be cooled  in a  desiccator and held  there
 until  they are  weighed.  If a desiccator  is  not  used,  the  sediment will
 accumulate ambient moisture and  the sample  weight will be  overestimated.
 A color-indicating desiccant  is recommended  so that spent  desiccant  can
 be detected eaily.   Also, the  seal on the desiccator should  be  checked
 periodically  and,  if necessary, the  ground  glass  rims should  be greased
 or the "0" rings should  be replaced.

                                    185

-------
                                                                  Sediment
                                                      Total Organic Carbon

     Calibration and Preventive  Maintenance—The  induction furnace used
for the total organic carbon analyses  should be inspected  and  calibrated
daily.   The analytical  balance and drying oven  should be  inspected and
calibrated weekly,  at a minimum.

     Quality Control Checks—Duplicate analyses  should be conducted on
at least 10 percent of  the  total number of samples.  An NBS  traceable  standard
reference material  should  be analyzed  daily to  monitor  performance.  A
method blank should be analyzed with each batch of samples.

     Corrective Action—If the  induction  furnace  does not appear to be
operating properly, follow  the manufacturer's  instructions for troubleshooting
and repair.

     Data  Quality and Reporting—Total organic carbon  should  be  reported
as a percentage of the dry weight of  the unacidified sediment sample to
the nearest 0.1 unit.   Results should be  reported  for all determinations,
including QA replicates.   Any factors that may have  influenced sample quality
should also be reported.
                                     186

-------
                                                                  Sediment
                                                  Biochemical Oxygen Demand
S'.ochemical  Oxygen  Demand (BOD)

Field Procedures--

     Con ection—Samples can be collected in glass or plastic  containers.

     Processing--BOD  samples should be analyzed immediately after  collection.
If  a  delay occurs,  samples should  be  stored at 40 C, and can be held  for
ua  to 7 days in that  condition.   Samples  should be kept  field  moist and
air contac'  should  be prevented to minimize oxidation.  Refrigerated  samples
should be v  irmed to 200 c prior to analysis.

Laboratory Procedures--

     Analytical Methods—BOD should be determined using the method described
by  Plumb (1981).  This procedure  is  similar  to  U.S. EPA  Method  405.1  and
APHA Method 507 for water and waste water and  specifies an incubation period
of  5 days at 20° C.

      It  is  critical that each  sample  be  thoroughly homogenized in the laboratory
before a subsample is taken  for analysis.   Laboratory homogenization should
be  conducted  even if samples were  homogenized in the field.   A subsample
should be analyzed separately  for total  solids so that  BOO can be determined
on  a dry-weight  basis.   The bacterial  seed  should be from a source having
a salinity  similar to that of  the environment from which  the samples were
taken.  The  salinity of the dilution water should  be  adjusted  with sea
salt (if necessary) so that  its salinity  approximates that  of the environment
from which the sample was taken.   Many  synthetic organic components in
sediments  are  not  biodegradable without  the  seeding  procedure because of
either  a toxic  effect or a deficiency or absence  of appropriate microorganisms.
Chlorine residuals must be removed  prior to the test because residual  chlorine
may be  toxic  to  the microbial  population or may  oxidi*e organic material.
 Because sediments  that contain  sulfide, sulfite,  or ferrous ions create
an immediate demand  on the dissolved oxygen, it is necessary to  distinguish

                                      187

-------
                                                                   Sediment
                                                  Biochemical  Oxygen Demand
this  immediate demand  from the true BOD.  The depletion of dissolved oxygen
during 15 min  in a standard water  dilution of  the  sample was  arbitrarily
selected by Plumb (1981)  as the initial oxygen demand.

     Calibration and Preventive  Maintenance—Dissolved oxygen measurements
should  be calibrated according to the procedures recommended in the appropriate
sections of this document.

     Quality Control Checks--Dup1icate  analyses should be conducted  on
a minimum of 10 percent of the total number of  :  mp.es.   A  dilution water
blank  and a g 1 ucose-glut ami c acid standard pro/ide quality control on  the
dilution water as. well  as on  the cleanliness of analytical  equipment (e.g.,
incubation bottles)  and should  each be analyzed  in triplicate with each
batch of samples.  The most reliable BOD determinations are made when residual
dissolved oxygen  is at least 2  mg/L and uptake of dissolved oxygen  is  at
least  2 mg/L after incubation (Plumb 1981).

     Corrective Action—Plumb  (1981)  and  APHA (1985) should be consulted
for methods of correcting for the many kinds of  interference that may accompany
BOD analyses.

      If the dilution water blanks  exceed 0.2 mg/L, cleanliness of containers
and water should be checked.   Containers may require  1+1  HC1 rinse after
detergent washing to remove and any residual  organic material.   Any containers
rinsed with  add should be thoroughly  rinsed with distilled water to prevent
acid  carryover.

      If  a 2-percent dilution of the  glucose-glutamic acid standard  check
solution  is  out sice the range of  218±11  mg/L, BOO determinations made with
the seed and  dilution water  are  suspect.  Several methods used to determine
the problem  include running a  series of dilution water  blanks using different
water sources with and without  seed,  preparing a  fresh solution of glucose-
glutamic acid,  changing the seed,  or preparing  fresh reagents  for the dilution
                                      188

-------
                                                                 Sediment
                                                 Biochemical Oxygen Demand
water.   The  source of the problem  should  be determined before additional
BOO analyses  are  performed.

     Data Quality and Reporting--BOD should  be reported  as mg/kg dry weight
of sediment to the nearest 0.1  unit.   The laboratory  should report  the
results  of all  determinations,  including  QA replicates,  dilution water
blanks, and glucose-glutamic  acid standards.  Any  factors  that may  have
influenced sample quality should  also be reported.
                                    189

-------
                                                                  Sediment
                                                     Chemical Oxygen Demand
Chemical Oxygen Demand  (COD)

Field Procedures--

     Col1ection--Samp1es  can  be collected in glass or plastic  containers.

     Processing--COD samples  should be analyzed immediately after collection.
If  a  del~.y occurs,  samples should  be stored at 4° C,  and can be held  for
up  to 7 -jays in that condition.   Samples must be kept field moist  and free
from air contact during stor  -,e to .rsinimize oxidation.

Laboratory Procedures—

     Analytical Methods--CQD  should be determined according to the method
described  in Plumb (1981).   This  open reflux method is similar  to  U.S.  EPA
Method  410.1  and  APHA Method 508A  for water and waste water  and consists
of  oxidizing the organic  matter  in a  reflux apparatus for 2 h.

      It  is critical that  each sample  be  thoroughly homogenized in the  laboratory
before  a subsample is taken for analysis.  Laboratory homogenization  should
be  conducted  even if samples were  homogenized  in the field.  A  subsample
should  be  analyzed separately for total  solids  so  that COD can be determined
on  a dry-weight basis.

      Because  traces of  organic material  from external  sources  may cause
a positive error,  care should be  taken  to avoid contamination of  glassware
and the distilled water used  to prepare reagents and to reflux the  sample.
Because volatile materials may be lost  when  the  sample  temperature  rises
during the  sulfuric acid  addition  step,  the  flask  should be cooled during
this  addition  to minimize this  loss.   Because  chlorides  are quantitatively
oxidized  by dichromate and represent a  positive  interference, mercuric
sulfate is added  to the digestion flask to  complex the chlorides.
                                      190

-------
                                                                 Sed imen t
                                                    Chemical Oxygen Demand
     Calibration  and  Preventive Maintenance--The technique  and  quality
of reagents should be evaluated  by  conducting the test on a standard potassium
hydrogen  phthalate solution.   Laboratory glassware should be kept very
clean to prevent introduction  of  organic material  to the sample.

     Quality Control  Checks--Duplicate  analyses should be  conducted on
a minimum of 10  percent of  the  total number  of  samples.   A method  blank
and  a  potassium hydrogen  ohi-halate standard should be analyzed with each
batch of samples.

     Corrective  Action—If results  of the analysis conducted on  the potassium
hydrogen phthalate standard indicate  that the  technique  is not  performing
properly, check glassware and the  distilled water for contamination, ensure
that reagents and  sample are thoroughly mixed, ensure that  excess  potassium
dichromate remains after  oxidation is complete, and critically  review all
other analytical steps.

     Data  Quality and Reporting--COD should be reported as mg/kg  dry weight
of sediment to  the nearest 0.1  unit.   The  laboratory  should  report  the
results of all  determinations,  including QA replicates,  method blanks,
and potassium hydrogen phthalate standards.  Any factors that may  have
influenced sample  quality should  also be reported.
                                    191

-------
                                                                 Sediment
                                                           Oil and Grease
Oil and Grease

Field Procedures--

     Col 1ection--Sediment samples  should be collected  In  wide-mouth glass
jars.  Bottles  should first be washed with a warm aqueous detergent mixture,
and  then,  in sequence, thoroughly rinsed with  hot  tap water, rinsed at
least twice  with distilled water, rinsed with l,l,2-trichloro-l,2,2-trifluoro-
ethane  (-',e.  preon or equivalent), and dried in a clean  oven at >_ 105° C
for 30 mir,   Bottle  lids  should be  lined with  TFE.  Aluminum-lined  lids
can  be  used, but contact with seawater corrodes  the  aluminum.  Plastic
containers  are not  acceptable.  Headspace  should  be left in the  sample
container for addition of acid and mixing or expansion during freezing.

     Processing—When  analysis  cannot  be  made within  24 h, preserve the
sample with approximately 1 ml of concentrated  hydrochloric acid  per 80 g
(wet weight) of sample.  Never preserve with chloroform or sodium benzoate
(APHA  1985).  Acid-preserved  samples  should  be stored at  40 C, and can
be held for up to 28 days  in that condition.   Although U.S. EPA has not
established a recommended maximum holding time for oil and grease in sediments,
28 days is  consistent with the recommended holding time for acid-preserved
water  samples.  Samples  can also be preserved by freezing at -20° C, and
can  be held  under that  condition  for up to 6 mo.   Samples must be kept
field  moist during storage because they may  lose apparent oil and grease
as a result of drying.

     A wet  sample  should be used for  analysis because a dried sample may
yield low results.   Concentration on a dry-weight basis  can be  obtained
by analyzing a separate aliquot for total  solids.

Laboratory Procedures—

     Analytical Procedures—Methods for  oil  and  grease analysis are described
 in Plumb (1981).  Soxhlet extraction of  the sediment  can be  inefficient
                                     192

-------
                                                                  Sed imen t
                                                            Oil and Grease

due to  channeling of solvent  through the sediment  layer.  Therefore, all
sediments should be stirred at least twice during the  extraction  period.
Because asphaltic materials are insoluole in  freon,  the recommended method
will give low recoveries for samples containing  these  materials.

     Determination of extracted  oil  and grease  concentrations by the gravimetric
method  is reliable for relatively nonvolatile  hydrocarbons, vegetable oils,
animal  fats, waxes, and  soaps.  Elemental sulfur  is co-extracted  with oil
and grease and will  interfere wit > gravimetric determinations.   Light  hydro-
carbons that volatilize at tempe-atures  below 700 C  (e.g., gasoline through
No. 2 fuel oil) are  lost during the solvent removal  step.

     Determination of extracted oil  and grease concentrations by the  infrared
method  may be more precise at lower concentrations than  is the  gravimetric
method.  This method will  give higher  results than  the  gravimetric method
if  the  sample contains  volatile components,  and will  give lower results
than  the gravimetric method if the  sample  is  high in  sulfur content.   A
limitation of the infrared determination  is  standardization,  which requires
a prepared  reference  oil.   The reference  oil may  not  be comparable to  the
type of oil  and  grease in the  samples, thereby resulting  in inaccurate
results.  When  the exact nature  of  the oil and grease in the  samples is
unknown, the reference  oil described in U.S.  EPA Method 413.2,  Section  6.4
 is  recommended.

      Calibration and Preventive Maintenance—For gravimetric oil  and grease
 analyses, check the accuracy  of  the analytical balance periodically (minimum
 of once per  week is recommended) using Class S weights.  A service contract
 that  includes  scheduled preventive  maintenance  at  least  once  per  year is
 recommended.   Scratched, chipped, or cracked  boiling flasks should be replaced.

      For infrared oil  and grease analyses, follow the manufacturer's preventive
 maintenance procedures  for  the infrared spectrophotometer.   Cells  used
  for analysis should be checked  for scratches  before each use.   Scratched
  cells  should not be used.
                                     193

-------
                                                                 Sediment
                                                            Oil and Grease
     Quality  Control  Checks--Repl icate analyses  should  be performed for
a minimum of 10 percent  cf  the total number of samples  to  establish an
estimate of precision.  Because spike results  may not reflect true extraction
efficiencies,  reference sample analysis,  if  available, is more appropriate.
A method blank should  accompany each batch of  samples.

                 Action--If concentrations of  oil  and grease in procedural
blanks are greater  than the detection limit,  check the cleanlin'ss  of all
glassware.  For infrared  analyses, severe  interferences will  result if
the freon solvent contacts any material  containing plasticizers (e.g, Tygon
tubing,  plastic bottles).   For gravimetric  analyses, high  results will
be obtained if Freon  or fumes remain in  the flask after distillation.

     Poor  precision may be  caused by  channeling of the solvent during ex-
traction.  To minimize channeling, ensure that  all moisture has been  removed
with the addition of  the magnesium sulfate and  stir the sample more frequently
during the extraction step.

     Data  Quality  and Reporting— The definition of oil and grease is  based
on the procedures  used.   Unless  identical  procedures are  used, oil and
grease determinations  are not intercomparable.  Thus, the method used for
analysis should always be specified and  consistent within a study.

     Reported values for precision and accuracy are  indicated below:

     •     Reproducibil ity for  gravimetric determinations of oil and
           grease in  sludge  samples  was 4.6-percent  coefficient  of
           variation (APHA 1985).  Variability  in sediment  oil  and
           grease determinations have often  been  reported as  high  as
           20-30 percent  (Disalvo et al .  1977).

-------
                                                                  Sediment
                                                            Oil and Grease
     t    Due  to the nonspecific nature of the  oil  and grease analysis,
          interpretation of spike  analyses  is  difficult.  Analysis
          of a reference material  (U.S.  EPA Municipal Digested Sludge)
          has  an acceptable recovery range  of  50-150 percent  of the
          reference value.

     Measurements should be reported to a minimum of  two  significant figures
and should clearly state whether concentrations are on a  wet- or dry-weight
basis.  Detection limits are in the  range of 1  mg/kg  (wet weight) for infrared
determinations nd 50 mg/kg (wet weight)  for gravimetric determinations.
Results  of all determinations should be reported, including QA replicates,
blanks, and reference samples.   Any  factors that may  have influenced  sample
quality should also be reported.
                                     195

-------
                                                                 Sed iment
                                         Sulfldes (Total  and  Water Soluble)
Sulfides (Total  and Water Soluble)
Field Procedures--
     Col lection--Samples  can  be  collected  in glass or plastic  containers.
Samples should be  removed from the grab  sampler as soon as  possible to
minimize exposure to air.

     Processing--Samples  should be analyzed for sulfides as soon as possible.
Samples for total sal fides  should  be preser  »d  oy adding  2N  zinc  acetate
solution (approximately 5 ml  for 30 g of seduent) and swirling the mixture.
Preserved  samples  for total sul fides  should  be stored in  the dark at 40
C and should be analyzed within 7 days.  Although U.S. EPA has not established
a recommended maximum holding time,  7 days  would be consistent with the
holding time recommended for acid-preserved water samples.   Samples  for
water-soluble sul fides  can be preserved by mixing 10-20 g of sediment in
a tared container containing 50 ml  of sulfur antioxidant buffer  (SAOB).
Preserved  samples  for water-soluble sulfides should be stored in  the dark
at 40 c and should be analyzed within 4 days. All preserved sulfides samples
should be stored in the dark.   It  is critical that air contact with  samples
be minimized and that samples be kept moist to minimize oxidation.

Laboratory Procedures--

     Analytical  Procedures--Total  sulfides should  be measured according
to procedures described in  Plumb  (1981).  A distillation  under acidified
conditions evolves hydrogen sulfide, which is swept  into a zinc  acetate
trap and precipitated as zinc  sulfide.  The sulfide  is then solubilized
and  measured using a colorimetric  procedure.

     Water-soluble  sulfides should  be measured  according  to the method
described  in Green and Schnitker  (1974).    The buffered sediment  must be
maintained in a slurry during  titration.  Although  zinc  or lead can be
used as a  titrant, cadmium produces the  sharpest  endpoint.  Because the

                                     196

-------
                                                                  Sed imen t
                                         Sulfides (Total  and Water  Soluble)
SAOB solution will oxidize and darken  with time, thereby losing its  buffering
capacity,  this  solution should be as  fresh as  possible.   Because  samples
high  in sul fides may exceed  the buffering capacity of the SAOB solution,
sample size  should be adjusted accordingly.

     For both  sulfides  techniques,  an  aliquot of  the  original  sediment
sample should be  analyzed for total  solids so that results can be expressed
on a dry-weight basis.

     Calibration  and  Preventive Maintenance—A series  f al least  three
standard sulfide  solutions should be prepare*1.   Because  si. 1 fide solutions
are  very  unstable, they should be prepared  fresh for each use  and used
iimiediately.  Stability of these solutions  can be  increased by using nitrogen-
saturated  water for dilution.  Because of variation among lots, the  methylene
blue solution should be standardized against a  standard  sulfide solution.
The  spectrophotometer used  for total  sul fides should be calibrated  daily
by zeroing  the instrument using a method blank and establishing a calibration
curve  using the  standard  sulfide solutions.   The curve  should  span the
expected range  of sulfides  concentrations.  The specific  ion probe  used
for water  soluble sul fides should be calibrated  according to the manufacturer's
instructions.

     Quality Control  Checks—Duplicate  analyses  should be conducted on
a minimum  of 10 percent of the total number of  samples.

     Corrective  Action—The amine-sulfurlc acid solution used in the color-
imetric determination should be  pink  initially and  turn colorless within
3 min  when  used   in the  procedure with a sulfide-free  sample.  If color
appears, prepare  a new stock solution with fresh materials.

     Data Quality and Reporting—A detection limit  of 5-10  mg/kg wet weight
can  be expected for both sul fides  techniques  using a 10-20 g  wet-weight
sample.  Sul fides  concentrations  should be reported as  mg/kg  of sediment
dry  weight to the nearest 0.1  unit.   Results  should be reported for all

                                     197

-------
                                                                     Sediment
                                          SuIfides  (Total  and  Water Soluble)

determinations, including QA replicates.  Any factors  that may have influenced

sample quality should also be reported.
                                       198

-------
                                                                 Sediment
                                                 Priority  Pollutant Metals

Priority  Pollutant Metals

Field Procedures--

     Co11ection--As with  water  samples, the best containers  for  collection
of sediment  for trace metal  analysis are made  of quartz or  TFE.  Because
these containers are expensive,  the preferred containers are made of poly-
propylene  or linear polyethylene  with a polyethylene cap  (APHA  1985).
Borr silicate glass containers  can be used and may be preferred if trace
orga. ic  compound analyses  are  to be performed  on the  same samples.   Do
not use  soft glass containers  or  containers with aluminum-lined or cardboard-
lined lids.

     Possible problems during  sample collection  involve contamination  from
the  sampling device, airborne  dust,  or cross-contamination  from previous
samples.   Contamination  can  be  minimized by avoiding the use of metal  when
collecting sediment samples.  If  metal must  be used, corrosive resistant
stainless steel  is the best  material.  When using a benthic grab or coring
device, contamination can  be minimized by removing only sediment that is
not  touching the  walls.  Prior to use,  sample containers  should be thoroughly
cleaned with a detergent solution, rinsed with  tap water,  soaked in acid,
and  then rinsed with metal-free water.   All glass or plastic parts associated
with the sampling  equipment  should be  cleaned  in  the  same manner.   For
quartz,  TFE, or glass containers, use 1+1  HN03, 1+1  HC1, or aqua regia
(3  parts concentrated HC1 + 1  part concentrated HNQ/j)  for soaking.   For
plastic material, use 1+1 HNQ.3 or 1+1 HC1.   Reliable soaking conditions
are 24  h at 700 C (APHA  1985).  Do not use  chromic acid for cleaning any
materials.   For metal parts, clean as stated  for  glass  or plastic,  except
omit the acid-soak step  of  the cleaning  procedure.   Acids used should be
at  least reagent grade.  If  trace organic compound analyses  are  to be performed
on  the  same samples, final  rinsing with acetone and then  high-purity methylene
 chloride  is acceptable.
                                     199

-------
                                                                 Sediment
                                                 Priority Pollutant Metals

     A minimum  sample size of 5  g  (wet weight) is required for  the analysis
of all priority  pollutant  metals.   To allow  for duplicates,  spikes, and
required reanalyses, a minimum sample  size of 50 g (wet weight) is reccrrmended.
To allow for mixing of the sample and  ease of  collection,  a 240-mL  (8-oz)
jar  is  recommended for collection.  A  125-mL (4-oz) jar would be adequate
but often difficult to fill.

     Processing--S.-.mp1es should be stored in clean containers after collection,
--i.d  picked in  ice while  in  the field.  Samples should be stored  at -20°
C   Although freezing is not required for  all  U.S. EPA  procedures,  it is
recommended to minimize  potential  alteration of analytes  by microbes.
Care should be taken to prevent container breakage during freezing.    Leave
sufficient headspace  for water to  expand  and  place the containers  at an
angle when freezing.

     No  recommended  holding  time for   sediments  has been established by
U.S. EPA.  A maximum  holding time of 6 mo (except  for mercury samples,
which should be  held  a  maximum of 30  days)  is  consistent with the  maximum
holding  time recommended by U.S. EPA for  water samples  (U.S. EPA 1985).

Laboratory Procedures--

      Analytical Procedures — Priority  pollutant metals should be analyzed
according  to procedures described  in Tetra Tech (1986a).   Prior  to removing
each aliquot for analysis, samples should be mixed thoroughly using nonmetallic
utensils.  Mix  all water  back  into the   sample.   If  there is  any question
regarding  nonrepresentative material   (e.g., twigs, leaves, shells,  rocks,
 and any material  larger  than 1/4-in),  U.S.  EPA should be contacted for
guidance.   A separate aliquot  should be analyzed for a total  solids  deter-
mination.

      Digest sediment  samples prior to analysis using the acids specified
 in the procedure  (Tetra Tech  1986a).   The digestate can  then be  analyzed
 by  flame  Atomic  Absorption  Spectrophotometry  (AAS), graphite furnace AAS,
                                     200

-------
                                                                 Sediment
                                                 Priority  Pollutant Metals

or  Inductively  Coupled Plasma (ICP), depending on the sample concentrations
and required detection limit.  Mercury digestion and analysis must  be  performed
on a separate sample aliquot by cold vapor AAS.

     ICP can  be used  to screen samples  for elements  that are present  in
relatively high  concentrations.  For those that may require  more sensitive
analysis,  graphite furnace AAS can be  used.  Analysis by ICP can be  subject
to  interelement  interferences, while  graphite  furnace AAS  can be subject
to  matrix  problems frcm  acid or  salt content of the samples.  Select  the
method with a detection  limit that  is adequate to determine compliance
with 301(h)  program criteria.

     Calibration and  Preventive Maintenance—In general, all instruments
must be calibrated daily and each time  the  instrument  is  set up.  For each
analysis,  calibration  procedures  should  follow those for the  specified
method.  Calibration standards must be  prepared  using the same concentrations
of  acids as will result  in the samples  following  sample preparation.

     After  an instrument has been calibrated, verify the accuracy of the
 initial cal ibration by the analysis of certified control  solutions at a
 frequency of once every  10 samples or every 2 h during  an analysis run,
whichever is more frequent, and after  the last analytical sample.   If a
certified control solution is not available, use  a  standard  that is composed
of the analyte from a different source  than  that for the  initial calibration.
 If the deviation of  the  continuing  calibration verification  is greater
than the calibration control  limits  specified 1n  the method,  the instrument
must be recalibrated,  and  the  preceding 10 samples  reanalyzed.

      All  equipment should  have scheduled  routine  preventive maintenance,
 and a record  of  all maintenance performed should be noted in a logbook.
 Critical  spare  parts should  be kept on hand.
                                     201

-------
                                                                Sediment
                                                Priority Pollutant Metals

     Quality Control Checks—Analyze standard  reference materials (SRM)
[e.g., the National Bureau of Standards  (NBS) Estuarine  Sediment  or  the
National  Research Council  of Canada  (NRC)  Marine Sediments]  to provide
a check  on digestion efficiency and  overall accuracy of the  analysis.
A minimum of one SRM should be analyzed  for each survey or 2  percent of
the total number of samples (i.e.,  1  per 50 samples), whichever  is more
frequent.

     To  estimate  r*ecision, 5 percent of the total  number of samples  should
be analyzed in  duplicate or one  duplicate for each survey,  whichever is
more frequent.  When more than 20 samples are to be  analyzed for one  survey,
the project manager may choose  to  implement a program of triplicate analyses.
The  overall percentage of replicates should be  at least 5 percent.  To
estimate recovery, analyze samples  spiked before digestion at the same
frequency as duplicates.  Add  spike  concentration approximately equal to
the concentration  found  in the  unspiked sample.   An acceptable range in
spike concentrations  is 0.5  to 5  times  the sample  concentrations.

     Carry a method  blank  through all  digestion  and  analysis steps  at  a
minimum  frequency of  once every 20 samples or once for each batch of  samples
analyzed, whichever is more frequent.   If the  concentration of the blank
 is  less  than the required detection  limit, no correction  of sample results
 is  performed.  If the blank contamination  is extensive (>30 percent of
 sample value) then the batch of  samples  associated with the blank  should
be  reanalyzed.   Data should be corrected by the data  user for the blank
 values between the required  detection  limit and the control  limit.

      For ICP analysis, additional  QC  checks  should include an interference
 check sample to  verify  interelement  and background  correction factors.
 For graphite  furnace AAS, additional  QC  checks  should include duplicate
 injections,  with the  mean value reported. The  relative  standard deviation
 of the  readings  should  be within  control  limits.  Otherwise, the sample
 should be reanalyzed.
                                     202

-------
                                                                  Sediment
                                                 Priority Pollutant Metals

     Corrective  Action—If the concentration of the field or method blank
is greater than the required detection limit,  all steps in the sample handling
snould  be reviewed.  Many trace  metal contamination problems are due to
airborne dust.  Keeping containers  closed and  rinsing all handling equipment
immediately prior to use minimizes  dust  problems.  In the field, mercury-filled
thermometers should be handled carefully or avoided because broken thermometers
are  a  potential  source of severe mercury contamination.  In the laboratory,
sanples  for mercury analysis should be isolated from items such as polarographs
or COD reagents.

     Poor  duplication may be  caused by inadequate mixing of  the sample
before taking aliquots, inconsistent contamination,  gross grain size differ-
ences, inconsistent digestion procedures, or instrumentation problems.

     Poor  performance on the  analysis of the  Standard Reference Material
(SRM) or poor spike  recovery may be  caused for the  same reasons  as  poor
duplication.  However,  if duplicate results  are acceptable, poor SRM performance
or poor spike recovery may be caused  by loss of analyte  during  analysis.
To  check for analyte  loss  during digestion and  for  low recovery due to
interferences during analysis, spike the sample after digestion and  compare
the  analysis to the  predigestion  spike.   If the  results  are different,
the  digestion technique should be adjusted.    If the results are  the same,
dilute  the sample by at least a factor of 5 and reanalyze.   If spike recovery
is  still poor, standard  additions,  matrix modifiers, or another method
is required.

     Data Quality  and Reporting—Report measurements as  mg/kg  to  a maximum
of  three  significant figures on a dry-weight  basis.  Detection  limits can
vary widely because of methods and  instrumentation.  Consult the  analytical
method  to  determine expected detection limits,  precision, and  accuracy.
Detection  limits actually obtained  should be  reported for  each  sample.

     The laboratory data summary  should include duplicate, spike, and  blank
results and  state clearly if and how any data were blank-corrected.   Data
                                     203

-------
                                                                   Sediment
                                                   Priority Pollutant Metals

to be  included  in  the ODES database  should  be  blank-corrected by the data
user.   The laboratory data suimary should also  include  the  following information
to allow independent QA review:

     •    Digestion procedures

     •    Quantity of sample digested and  final  dilution  volume

     -i    Percent soi ids

     •    Instrument detection  limit for each  element

     0    Method of detection  (I.e.,  graphite furnace, flame,  ICP,
          hydride, cold vapor)

     •    Deviation  from the prescribed methods

     •    Blank  associated with sample

     •    Problems associated with analysis.

For  a more  thorough QA review, additional documentation (e.g., calibration
curves) may  be requested.
                                      204

-------
                                                                 Sediment
                                      Priority Pollutant  Organic Compounds

Priority Pollutant Organic Compounds

Field Procedures--

     Col lection--Priority  pollutant organic  compounds  can  be separated
into  purgeable  (volatile)  and  extractable (acid,  base, and  neutral,  including
pesticides and PCBs) compounds.  Container preparation and collection techniques
differ for these two groups.   Sediment samples for analysis of  extractable
compounds should  be collected  in 240-mL  (8-oz) or larger, wir^e-mouth glass
jars with TFE-lined screw lids.  The container,  lid,  and  lii.jr should be
detergent washed, rinsed twice with tap water, orce with distilled water,
once with acetone, and once  with high-purity methylene chloride.   Firing
of the  glass  jar at 450o C  for  1 h may be substituted  for the  final  solvent
rinse.  Collection of a minimum  of 200 g (wet  weight)  of sample  should
provide enough  material for a full analysis and all  required  QC  analyses.
Headspace should be left to  facilitate mixing of the sample.

      If a volatile analysis  is required, two separate 40-mL glass containers
should be filled leaving no headspace.  This container,  screw cap, and
septum  should be washed with detergent,  rinsed once with tap  water, rinsed
again with distilled water, and dried at >105° C.  Use of a  solvent  rinse
will  interfere with the analysis.   To obtain a sample with no headspace,
fill  the vial to overflowing  so  that a convex  meniscus forms at  the top
if there  is adequate water  in  the sediment.   With the  liner's TFE  side
down, place  the cap carefully on  the opening of  the vial, displacing the
excess water.  Once sealed,  invert  the bottle to verify the seal by demon-
strating the absence of air bubbles.   Samples for  volatile analyses should
be taken only  from  single grab samples  because many of the  volatile compounds
of  interest may  be lost while compositing.

      An empty sample jar carried  through all  processing and  handling will
 serve as  a field blank.  A solvent rinse of  the bottle  should  be analyzed
                                     205

-------
                                                                  Sed imen t
                                       Priority Pollutant Organic  Compounds

by the  lab with each batch of samples  received to serve as a check on con-
tamination that may occur during shipping  and  storage.  The  solvent should
be the same as that used for sample extraction.

     Process ing--Samples should  be stored  in the  dark at 40 C, on ice,
or frozen until extraction.  Care should  be taken with  frozen samples  to
prevent container breakage by leaving  headspace for the water  to  expand
and by freezing containers at an angle.   U.S. SPA   gives no official guidance
on  sediment holding  times but recommends  that water  saroles for extractable
organic compounds stored at 4° C be extracted within  l«l days of collection.
Because sediments can  be  frozen at -20°  C, longer holding times  (e.g.,
up to 6 mo) are appropriate.  Extracts  should be analyzed within 40 days
of  extraction.   Effort should be made  to analyze  the samples as soon as
possible after extraction because  some  of the  more labile analytes may
degrade  in solution.   Degradation may occur  even  in  the dark under  refriger-
ation,  possibly as the result of free radical   formation.   Freezing  is not
recommended  for  volatile samples  because no headspace  is to be left  in
the  vials.   A maximum holding time of 14  days  for  analysis of volatile
compounds in  sediment  samples would  be consistent with the holding time
recommended by  U.S.  EPA  for  water samples.

Laboratory Procedures--

      Analytical  Procedures--Priority pollutant organic compounds should
be determined according to procedures described  in Tetra Tech  (1986a).
 Before  each  aliquot 1s removed for analysis, the sediment and any standing
water should  be stirred well  to  create a  homogeneous mixture.   If there
 is any  question  regarding  possible  nonrepresentative material  (e.g., twigs,
 leaves, shells, rocks, and any  material  larger than l/4-in), request guidance
 from U.S.  EPA.  A separate aliquot  should  be  analyzed for a total solids
 determination.

      To obtain meaningful  results,  an  efficient  extraction of the compounds
 of interest must be performed.  The  length of  time a sample should be extracted
                                     206

-------
                                                                 Sed imen t
                                      Priority  Pollutant Organic Compounds

varies with  method  and  matrix.  When dealing  with an unfamiliar  method
or matrix,  step-by-step testing should be performed  to deterrr.lne completeness
of extraction  and  to identify special  problems.   For instance, it has  been
found when using the Soxhlet method  with  wet sediments,  the sample must
be stirred two  to three times during the procedure  to prevent channeling.

     Removal of elemental  sulfur  from marine  sediments is  critical  for
improving column resolution in gas chromatography,  and  for low-level detection
of compounds such  as PCBs by gas chromatography/electron capture.  Cleanup
of the  ample extracts using column  chromatography  is  often necessary  to
remove  interferences.  Calibration of the columns enables the compounds
of interest to  be collected  in the proper fraction.

     Cross-contamination between  samples  should be avoided in all steps
of analysis beginning with clean glassware.  Injection micro-syringes must
be cleaned well  between uses.  If separate syringes are used  for injection
of standard solutions any bias between  syringes should  be accounted  for.
Carryover can occur when  high- and low-level samples are  analyzed sequentially.
Analysis of an appropriate solvent blank  following a  high-level  sample
may be necessary to check for carryover.

     Generally, achievable detection limits for sediments  are outlined
by specific methods.  Actual detection limits  vary  depending  on  sample
size,  final volume, co-extractive compounds, and  the  source  and nature
of sediment.  Select  the method  such that  detection limits  are adequate
to determine compliance  to 301 (h) program  objectives.

     Calibration  and Preventive Maintenance—Before beginning analysis
of samples, a calibration curve that  brackets  the working  range must be
established.  This  calibration  should be repeated  after each major equipment
disruption.  Calibration checks  of  the  GC/MS system should  be done at the
beginning and end of the day  and at least  every 12 h to demonstrate that
the  instrument's response is  within  control  limits.  Specific tuning  criteria
 (e.g., OFTPP, BFB)  are provided  in  each  method.  Calibration checks  of
                                    207

-------
                                                                 Sediment
                                       Priority Pollutant Organic Compounds

the GC/ECO system should be  done at  the beginning of the day and verified
at least every 6  h.  These are only recommended minimum  frequencies  and
the nature of the samples may necessitate more  frequent verification.

     A routine QC check for each lot  of the analytical reagents being used
in extraction can prevent undetected contamination problems.  Also,  each
lot of alumina,  silica  gel,  sodium sulfate,  Florisil, resins, or charcoal
used should be monitored as a possible  source of  contamination and  cleaned
as necessary.   Each  lot of surroga e  mixture  should be checked for contam-
inants.  The  efficacy of the mater als  listed needs to be  evaluated as
there  is variation between lots.  Proper  storage is essential for these
materials.

     Equipment should be maintained  and serviced  routinely according to
manufacturer's instructions  and good laboratory practices.  Logbook  records
should document maintenance for  each measurement device.  Critical spare
parts  should  be kept on hand.

     Quality Control  Checks—Duplicates and matrix  spike analyses  should
be performed  to measure precision  and accuracy.  A  frequency  of  5 percent
of each  or one each per survey, whichever is more frequent, is  the reccmnended
minimum.   When more than 20  samples  are to  be analyzed for one survey,
the project manager may choose to  implement a program  of  triplicate  analyses.
The  overall percentage of replicates should be at least 5 percent.  Method
interferences can be caused  by  contaminated glassware,  reagents, solvents,
or processing hardware.  These materials can be monitored for contamination
by processing of 5 percent method blanks, or one per sample  set, whichever
 is more frequent.  Addition  of  known amounts  of surrogate compounds  to
each  sample will serve to monitor preparation  and  analysis  of samples.
 Standard reference material should be analyzed if available, as another
measure of accuracy.  Once  for  each survey or 2 percent  of the total  number
 of samples, whichever  is more  frequent, is the recommended minimum.
                                     208

-------
                                                                  Sediment
                                      Priority Pollutant Organic Compounds

     Corrective  Action--Uhen  results of QC samples fall outside of established
limits, several courses of  action are available.   Contamina :ion in the
lab reagent blank sample is cause  for  similar findings of the  same compounds
in samples to be suspect.  If  contamination  is  extensive,  reanalysis of
the whole associated group may  be  in order.   Blank contamination  should
be kept to less  than 10 percent of sample values  and preferably below the
method detection limit.  Contamination found  in  the field blank  should
be considered when looking  at the  associated samp'ie data.  Extensive contam-
ination of lab or field Hanks  (>30 percent of sample values)  should lead
to a detailed review of labi.ratory, sampling,  transport,  and storage pro-
cedures.   Phthalates, methylene chloride, and toluene are common laboratory
contaminants that may be detected in blanks  above the  method detection
1imit.

     Poor duplication may  be caused by  inadequate  mixing  of the  sample
before removing aliquots, inconsistent  contamination, inconsistent extraction
procedures, or instrument problems. The project manager should be  contacted
when results exceed the control limits.  Further replication of an  analysis
may be necessary to determine the reason for the poor results.

      Poor spike recovery may  be caused for  the  same reasons as  poor duplication
or by  matrix effects  produced by  co-extracted materials.   If the  spiked
compound  is added  at a concentration much  less  than  that  found in the  sample,
 recovery  may be difficult to determine.  This  problem  is difficult  to  avoid
 as most  environmental  samples contain  unknown concentrations of organic
 compounds.  To check  for analyte  loss during processing, a step-by-step
 examination of the method using a  spiked  blank  is  necessary, with measurements
 of the analyte at each  step.

      Sample results that fall  outside the established  calibration curve
 are  suspect  until  linearity of  the  instrument  response can  be  shown  at
 that  concentration  or until  the extract is diluted appropriately and re-
 analyzed.  Extremely high concentrations of organic  compounds may  saturate

                                     209

-------
                                                                  Sediment
                                       Priority  Pollutant Organic  Compounds

the  extraction capabilities of the method  and  may necessitate re-extraction
of a smaller sample  size or use of a more appropriate method.

     Emulsions,  colored extracts, or unusual  chemical behavior of a sample
should be noted and  considered when  reviewing results.   Modifications  of
a method by an experienced chemist may alleviate  some problems.  All  deviations
from specified methods  should be documented in  logbooks.

     If  the continuing calibration (single  point)  falls outside control
limits,  no  sample' should be analyzed until the calibration  is within  these
limits.  The  standard should be reinjected  to confirm  the problem and to
discount the possibility of operator error.   If still  outside of  control
limits, the  instrument  should be  recalibrated  (multi-point), and at least
the  previous sample reanalyzed and  results compared.   This may  indicate
that reanalysis of all  samples since last calibration  is  unecessary.

     Data  Quality  and Reporting--A data  summary for each  sample should
be  submitted.  All data should be dry-weight corrected, and reported as
ug/kg using  two  significant figures.  Data should not  be  blank-corrected.
Spike  recoveries, relative  percent difference between duplicates,  and  blank
results (ng/sample)should also  be submitted.  The  following information
 is  also needed for each sample  to  allow  independent QA review:

      t    Sample weight extracted

      •    Percent solids

      •    Final volume of extract

      •    Amount  of extract injected

      •     Instrument detection limits

      •    Method  detection limit  for each compound
                                     210

-------
                                                                    Sed iment
                                        Priority Pollutant Organic Compounds
     •    Blank associated  with sample


     •    Deviations from the prescribed  method


     •    Problems associated with analysis.


For a more thorough QA review, additional documentation (e.g., chrcmatograms,

computer listings) may be requested.
                                      211

-------
                                                                 Sed Imen t
                                                                  Infauna
Infauna

Field Procedures--

     Construction of Sieve Boxes—Sieve boxes should be of sturdy construction,
with high sides to minimize  the  possibility  of  material washing  out of
the  box.   They should also be large  enough to receive the  benthic sample
and wash water without clogging.   Swartz  (1978)  recommends  boxes 40 cm
x 40 cm.   The boxes  should also be constructed to  permit  nesting of -he
sieves, especially if more  than  one mesh  size will  be  used.   A  typical
sieve  box might be constructed  as in Figure  8.  Note the  application of
silicone sealant at the mesh  wood  interface.   This  sealant  will  prevent
organisms from crawling into the  space where the mesh  enters  the  box frame.
All wood pieces used in construction of the  sieve  boxes should be  treated
with fiberglass or epoxy resin (of the  types used in boat building),  sanded,
and painted.

     It is imperative that the mesh  used  in  the  sieve boxes meet specifications
outlined in ASTM E-ll, USA Standard Z23.1, AASHO M92, and  Fed. Spec.  RR-S-366b.
Such mesh is  available  from  scientific supply houses.  Inferior mesh will
not  have uniform openings and will not  be durable.

     Before each cruise, the sieves should  be examined for damage and wear.
Look for rips in the mesh, irregular mesh spacing, and sand  grains caught
in  the mesh.  Use water pressure  to  dislodge the sand.  Do not use sharp
objects, as the mesh may be damaged  or  the mesh  spacing may be altered.

     Fixative Preparation—The  fixative  most commonly used for  benthic
infauna samples is formalin,  an aqueous  solution  of formaldehyde gas.
Under  no circumstances  should  ethyl  or  isopropyl alcohol (preservatives)
be  used  in place of the  formalin.  Penetration of the  alcohol  into body
tissues  is too slow to prevent decomposition of  the specimens.
                                    212

-------
                                  rXftHitt*tt*Mi%i%*f
                                  f^ilr^fii^afM^m^^m^mlfm-fmft^tflf
•SCREEN LAPS OVER BOTTOM SIOEP1ECE
                             CONSTRUCTION
                 1. Construct upper and lowwr box frames (A.B).
                 2. Nail or stapt* mMh ovw lower tram*.
                 a Mount upper frame on lower frame.
                 4. Nail and glue side pieces (C) on upper and
                   lower frames (A.B). offsetting to permit
                   sieves to be nested.
                 & Use waterproof glue (reaortinol. epoxy) and
                   nails throughout construction.
Figure 8.   Construction of a  sieve  box.
                                   213

-------
                                                                  Sed iment
                                                                   Infauna

     Formalin  solutions of 5-20 percent strength  are  recommended for fixing
marine organisms (Gosner 1971; Birkett and Mclntyre  1971; Smith and Carlton
1975;  Swartz 1978).   Solutions of  10-15  percent are used most commonly.
It is recommended  that at least 2  L of diluted formalin solution be  on
hand  for each replicate sample to be collected, unless experience has shown
otherwise.

     The  formalin solution  should always  be  buffered to reduce acidity.
Failure to buffer may result  in decalcification of mr  '.-^scs and echinoderms.
Ideally, pH should  be at least 8.2, as calcium carbciate  dissolves in more
acidic  solutions.  Borax (sodium borate, Na£B407)  should be used as a buffer
because other buffering agents may hinder  identification by  leaving a precip-
itate on  body tissues and  setae.

     To prepare a 10-percent buffered  formalin solution,  add 4 oz of borax
to each gallon of concentrated  formalin (i.e.,  a  40-percent  solution of
formaldehyde  in water).   This amount will be in  excess, so use the clear
supernatant  when making  seawater  dilutions.   Dilute the concentrate to
a  ratio  of  one  part concentrated formalin to nine parts  seawater.   Seawater
will  further buffer the solution.   Seawater  also makes the  fixative  isotonic
with the  tissues of the animals, thereby decreasing the potential  for animal
tissues to swell and break apart, as often happens with freshwater dilutions
of formalin.

      It is desirable to prepare fresh fixative prior to each sampling excursion,
as formalin  will eventually  consume all the  buffering capacity of the borax.
 This occurs  because  the formaldehyde gas is  in  polymer form when in  solution.
 As the borax neutralizes the excess  acidity, the  polymer  chains become
 unstable and break apart,  exposing new acid sites.   Do not expose  formalin
 solution of any  strength to cold  temperatures because  the formaldehyde
 polymers will degrade into paraformaldehyde and  the  solution will  have
 to be discarded.
                                     214

-------
                                                                 Sediment
                                                                  Infauna
     Rose Bengal Preparat'on--Rose bengal may be added  to the  fixed samples
either as a powder or a solution.   Both are  effective.  However,  it is
easier, and  perhaps less  expensive, to use a solution.  Simply  mix  a concen-
trated solution  of rose bengal  in  tap water and put it in a dropper bottle.
The  relative quantity of stain  being used is much easier to  estimate  when
dispensed from a dropper  than when added  in powder form.

     Relaxant Preparation—Several relaxants may be used  on benthic organisms
prior to -ixation.  However, a  solution of magnesium chloride  in  tap water
*s effective on  a wide variety  of  taxa (Gosner  1971), and is easily prepared
and  used.  The MgCl2 solution  should  be  isotonic with seawater.  To prepare,
dissolve 73-g MgCl2*6H20 per  liter of tap water.  Anhydrous MgCl?  can be
purchased (at considerably more  cost)  and  used to prepare  the relaxant
solution.   However, accurate  determinations of mass are very  difficult
because of the propensity of the crystals to  absorb atmospheric moisture.
Hence, use of the hydra ted form is recommended.

     Sample Containers—It is recommended that  glass or plastic  jars be
used for sample  fixation and storage.   Plastic  lids are  preferable to metal
lids because formalin  corrodes  metal.  One- or two-quart containers are
usually adequate for a 0.1-m2  sample after washing.   However, more or larger
(i.e., gallon) containers may  be  required  if large  quantities of gravel,
peat,  wood chips, or other large  particles  occur  in  the sample.  All  sample
containers  should  be labeled  internally and externally using waterproof
materials.

     Screen Mesh  Selection—It  is critical  that an  appropriate screen mesh
size be  chosen for separating the  infauna from  the sediments.  The  most
frequently used sieve  mesh sizes have been 0.5 mm and  1.0 mm.  These  sizes
roughly  correspond  to a  trough  in the size spectrum for  benthic organisms.
This trough  occurs at  0.5-1.0  mm  and  separates meiofauna from  macrofauna
 (Schwinghaner 1981).   Therefore, major deviations from these (or intermediate)
screen  sizes  will  greatly affect  the suite of organisms  collected.
                                     215

-------
                                                                 Sediment
                                                                  Infauna
     Mesh  size  selection  depends largely on  the objectives of the study.
Macrofaunal recruitment ana settlement studies definitely  require the use
of 0.5-mm mesh,  while studies  of the distributions and abundances of primarily
adult animals  usually require only  1.0-mm mesh.  The major disadvantage
of  using  the  larger mesh size is  that  some small species,  particularly
polychaetous  Annelids, may be lost.   Disadvantages associated with use
of  the  smaller mesh size are primarily related to processing time.  The
residual samples (after washing) v  >ntain more detritus and sediment,  more
juvenile  macrofau.ial forms,  and more  meiofauna.   Therefore, they  require
considerably more time to sort.   The subadult forms are also much  more
difficult to  identify, which increases the identification  time.   It is
not unusual for  samples washed on  0.5-mm mesh to require  50 percent  more
processing time than sampl.es washed on 1.0-mm mesh.  One may argue  that
the retention of juveniles permits a more complete characterization of
the  community.  However, because  many  of the juveniles  that recruit  into
a community are rapidly removed through competition or  predation, their
inclusion in  samples may not represent the more  stable adult community
that  eventually inhabits a site.

      A  wise practice 1s  to use a set of stacked screens with a 1.0-mm mesh
sieve sitting  on top of a 0.5-mm mesh sieve.   Both  subsets of the sample
can be handled  separately,  and  either or both can be examined based  upon
study objectives. Monitoring  programs conducted on  the  west coast of the
United  States generally rely on the  1.0-mm analysis,  while  the 0.5-mm sample
is  sometimes archived in  case something unusual  is found.  By  contrast,
studies on the  east coast of the United States tend to  sample  the 0.5-mm
fraction more frequently.

      The  same mesh size  should be used  consistently through time to permit
temporal  comparisons of  the  biota within  the study area.   If  possible,
the  chosen mesh size  should be  consistent with that used for  past studies
 in the  area and  for other  studies within the biogeographic region.   Such
                                     216

-------
                                                                 Sed inent
                                                                  Infauna

consistency  w 11  facilitate biological  comparisons with  historical data
and with data  from adjacent  areas.

     Collection--Collect benthic samples  in accordance with the on-board
procedures  for grab sampling discussed earlier.

     Proc>iSsing--After  qualitative characteristics of the sample have been
recorded,  sediments should  be  washed on the  designated sieve(s),  using
one of  several  possible me» tods.   Sediments may  be gently sprayed with
water from  above, agitated by hand  in a washtub of water, or washed  using
a combination of these techniques.  For all methods, it is  imperative that
the samples be washed gently to  minimize specimen damage.  A few  minutes
extra care in the field can save hours of  time for the  taxonomist, and
will result in a better data set.

     For most surveys,  it  is probably  easiest  to  wash the  samples from
above with  a  gentle spray, because  efficient, easy-to-use  gear may  be con-
structed  to  hold the grab  and  sieve boxes.   An example of a grab stand
is shown in  Figure 9.  The top section  is designed to accept the grab  sampler.
Wash water and  sediment drain  through the  openings in the bottom  of the
top tray and  into the lower  section of the sieving stand,  where  the screen
box(es)  is  (are)  located.

     All wash water should  be filtered or screened through mesh with openings
one-half the size of those  used in the survey, so  as not  to introduce plank tonic
or  bentho-pelagic organisms  into the samples.  Failure to screen  in this
^ay can  result in  increased  sorting  time.  It can also compromise  the  quality
of  the  samples, because it is  impossible to distinguish bentho-pelagic
organisms caught  by the grab from those  entrained  in  the wash water.

     Sieving  stands should  be designed  to  fit the specific sampling vessel,
and should never be  too large for the  available  space.  Moreover, they
should  have attachment points (e.g.,  eyebolts)  at  appropriate  places with
                                    217

-------
                                                         SPOUT
                                          SIEVE TRAY



                                       EYE SOU-
                              REFERENCE: STRIPLIN AND HAUPIN (1982)
Figure 9.  Example of a sieving stand.  Screen boxes
           (not  shown) are  placed in sieve tray.
                            218

-------
                                                                  Sediment
                                                                   Infauna

wh-ch  the  stand may be Mshed  to the deck or rail.  As shown  in  Figure 9,
all waste water  should  exit the  sieve tray  via  a spout,  to which a hose
should be attached.  The wash  water can then be discharged overboard through
a scupper.   This is  especially important  in cold weather, when  wash waters
may otherwise freeze on the deck  and  safety may be compromised.

     After  the  contents cf nne grab  are emptied  into the sieving stand,
the sampler should be washed of all  sediments adhering  to the inside  and
prepared for descent.  Take  care not to wash a  y  sediment clinging to  the
outside of  the  grab into  the sample.   Sediments  in  the  top  tray  should
be  broken apart  gently using the spray.  Breaking  up the sediments in  the
upper tray and not  1n the screen boxes  eliminates one  source by which infaunal
organisms become damaged.   Sediments retained  on  the  sieve screens should
be  gently washed  in the lower section of  the  sieving stand.

      Once  sieving  is  completed, the screen box should  be held at an angle
 and the  remaining material gently washed  into one corner.   The sample  may
 then be transferred  to an enamel  pan  or  jar for  relaxation, or to a jar
 for fixation, using as little water as  possible.  Place the permanent  internal
 sample label  in the enamel  pan or jar at this time to avoid  confusion later.
 If more  than  one  screen fraction 1s generated, be sure to keep  them separate
 throughout all phases of  field  and laboratory processing.   Be  sure to check
 the  screen for organisms trapped  in  (or  wound around)  the mesh  wires.
 If they cannot  be dislodged with gentle water pressure, use a pair  of jewelers
 forceps.  Be careful not to  damage  the  wire mesh.  After the  screen has
 been  checked for remaining animals and  sample removal is complete, backwash
 the screen with a high pressure  spray to  dislodge any sediment grains  that
 may be caught in  the mesh.  Repeat these  steps after every sample.

      Although  relaxation of the organisms in  the  sample is often omitted
 from  field sampling programs, 1t is  highly recommended  for  two reasons.
 First,  relaxed organisms are less  likely to distort their shape when fixed,
 because the  muscles  are  less  likely to contract.   The  more natural  the

                                     219

-------
                                                                 Sediment
                                                                  Infauna

appearance of the body, the  easier  it is to identify the animal.   Second,
relaxation helps  prevent autotomy,  the process whereby organisms (especially
polychaetes)  fragment themselves in reaction  to  the  fixative.   Complete
animals are much  easier to identify than fragments,  and  are necessary  for
the  identification of species  in some taxonomic  groups  (i.e.,  maldanid
polychaetes).

     Once  the  sample  is in a jar  or enamel pan, it  should be ii-wersed in
relaxant (e.g., MgCl2) for about  30 min.  After immersion, the sample should
be poured through a small sieve with mesh openings hair  the size of those
required for washing the  sample.  Be  sure  to  recapture  the relaxant, as
it may be reused  several  times  (Fauchald  1977).  After the animals  are
screened, gently transfer the sample  to a jar for fixation, using as little
water as possible.

     In  lieu  of  using Mgd2  as a relaxant,  it may be possible to minimize
damage to the organisms by increasing  the  strength of the formalin  solution
(Word, J., 13 February  1985, personal communication).  A strong  solution
may kill the organisms  before  they have  time  to contract or fragment.
Some experimentation may  be needed to determine whether this method is
effective  in a given region, and  if so,  at what formalin  concentration.

     As  mentioned earlier, a  10-percent solution of borax-buffered  formalin
 is  usually  used  to fix benthic organisms.  However,  samples containing
 large amounts of fine-grained  sediments, peat, or woody plant  material
 require  higher concentrations.  The sample should be covered with  fixative
 to  a mininum  depth of approximately  one  third that of the  sample itself.
 Thus,  the largest sample  that  can  be placed  in a  jar occupies  only three-
 quarters  of  the available volume.  The fixative will  occupy the  remaining
 one-quarter of the volume.  The  formalin  solution  should  be  added to the
 sample  jar until it  is completely  filled.  This will minimize  abrasion
 during shipping  and handling.  If  the sample volume exceeds three-quarters
 of the jar volume, more than  one jar should be used.
                                     220

-------
                                                                 Sed iment
                                                                  Infauna
     Rose bengal may  now be added to  the  sample.   Add  enough to turn the
fixative cherry-red.  This vital  stain aids the  sorting process by making
animals  more visible.  However,  it renders the body  of the organism opaque,
and thereby makes some taxonomic  manipulations difficult.  For this reason,
the addition of rose bengal  is  considered optional.

     After  the formalin  end rose bengal  (if desired)  have been added to
che sample jar, close the container and  gently mix the contents  f.   ensu.-e
that  adequate fixative  reaches all  the organisms.   The  samples  should now
be placed in protective containers for storage and  transport  to  the laboratory.
Wooden  boxes with handles  are  good  containers  for jars because they are
durable and easily constructed.   Bags should be placed in durable water-tight
containers such as  snap-lid buckets.

     On  board ship,  samples  should  be stored so  as  to minimize exposure
to  sunlight  and temperature extremes.  They  should  also  be  stored in  a
stable part  of the ship to minimize  agitation.   If special  racks are not
available to  keep  the  shipping boxes  stationary,  they should  be  lashed
down.

Laboratory Procedures—

      Equipment and Supplies—The  laboratory  should  be equipped with both
 stereo dissection  and compound microscopes.  One dissection microscope
 capable of  magnification  to  25-power should  be  available  for each  sorter,
 and one dissection microscope capable of magnification to  50-power  should
 be available for  each taxonomist.  Compound  microscopes should be  capable
 of magnifications up to 1,000-power.   At a minimum,  one compound microscope
 for each  three taxonomists is needed;  ore compound microscope for each
 two taxonomists is preferable.   The optics of the dissection and compound
 microscopes should  be of  high  quality to reduce sorting and identification
 times and  to increase acccuracy  by reducing the chances for misidentifying
                                     221

-------
                                                                 Sediment
                                                                  Infauna

organisms.   Other necessary laboratory supplies  include jewelers forceps,
fine scissors,  small scalpels, fine needles, flat and depression microscope
slides,  cover  slips, small dissection trays, immersion oil,  and glycerol
alcohol (half glycerol and half 70-percent alcohol).

     Preservative  Preparation—After the specimens  are  fixed, alcohol  should
be used as a long-term preservative.   Either 70-percent  ethanol  in water
or 70-percent isopropanol may be used.  Although isopropanol  is  less expensive
than ethanol,  it  is more unpleasant to work  wit  .   Specimens preserved
in  isopropanol are unsuitaole for  histological examination.  If future
studies of anatomy or reproductive  biology are  anticipated, ethanol must
be used.

     It  is  most cost-effective to purchase isopropanol and  ethanol in  bulk
solutions of 5-percent water and  95-percent  alcohol.   Purer grades  are
available, but more costly.  To  prepare 1 L  of a 70-percent solution of
either  alcohol,  add 263 ri. of water  to 737 ml of 95-percent alcohol solution.
It  may be necessary to use distilled water to dilute the  alcohol solution
because hard water mixed with  alcohol creates a milky  precipitate  that
makes  examination of the   samples difficult.  The alcohol  solution is  most
easily stored in a glass jar with  a clamped  siphon  hose.  This apparatus
allows workers to  accurately dispense small quantities  without  wastage.

     Use  of the 70-percent alcohol /30-percent water solution  is adequate
for  the preservation  of most infaunal organisms.   For long-term storage
of  crustaceans,  however,  it is  recommended  that glycerine be  substituted
for  some  of the water.  The  glycerine helps  keep the exoskeletons  supple,
thereby  facilitating examination and manipulation.  This  is especially
critical  for crustaceans archived  in  the reference collection  (see below).
An  appropriate glycerine-alcohol solution would be 5-percent glycerin in
70-percent  alcohol.
                                     222

-------
                                                                 Sediment
                                                                  Infauna
     Analytical Procedures--Samp1es  should remain in the formalin-seawater
solution for  a minimum of 24 h to  allow proper  fixation  (Fauchald 1977).
After  24  h,  the samples may  be  washed (i.e., rescreened)  on  a sieve with
mesh openings half the size of those used in the field.   The  smaller screen
size ensures that specimens  collected in the  field  will  be  retained  in
the sample regardless of shrinkage or  breakage resulting  from contact with
the  formalin.   It is desirable  to  wash the  formalin from the samples  as
soon as possible after  the  initial  24 h because the buffering capacity
of the bort/  in  the formalin solution  decreases continually.

     If the  sample  consists of  multiple  jars,  locate  all jars  prior  to
rescreening and  wash  them at the  same  time.   Carefully  pour the contents
of each jar into  the appropriately sized screen and  rinse the jar to remove
adhering organic  material, sediment, or organisms.  To reduce  the possibility
of sample loss through  spilling or  splashing, do not fill  the screen more
than half full.

     Caution should  be exercised when handling formalin mixtures because
formalin  is  toxic  and carcinogenic (Kitchens et al. 1976).  It can cause
irritation to the  eyes,  nose, and throat at  concentrations as low as 1.0 ppm.
Sensitivity in humans varies with the individual,  but in general, the detection
limit  is approximately 2 ppm.   The  technician doing the  rescreening should
wear protective clothing, rubber gloves,  and safety  goggles, and  should
work under  a properly  ventilated  fume hood.  A protective vapor mask should
be worn,  even when working near open windows or under a ventilation hood.

     There are  several  acceptable methods for rinsing formalin  from a sample.
One  method  is to gently flush the sample  with  large  quantities  of fresh
water  from  a low-pressure faucet or hose, being careful not  to splash  any
 sample material.   The rinse  water  should be  discharged  into  a  sanitary
 sewer.   A  second method is to  partly immerse  the sieve  in  a plastic  tub
 filled with  fresh water and wash  the  sample by  moving the  sieve  in an   up-
                                     223

-------
                                                                 Sed inen t
                                                                  Infauna
and-down  motion.  Care nust be taken  not to let the  water rise above the
top of the sieve.

     Allow  the  rinse  water to completely drain from the  sieve and lightly
rinse the sample with a solution  of  70-percent ethanol from a squirt bottle.
Carefully wash the sample material  into a  sample  jar filling  it no more
than three-quarters full.   Avoid  scraping  the  sample across the  screen
with  the spoon.  Rinse the  i*st  bit of material  into the jar using the
squirt bottle of alcohol.   F  1  tne jar  to  the top with the 70-percent
alcohol  solution and  screw  tne  lid on tightly.   Gently shake  and  invert
the jar several  times  to ensure proper mixing.

     Each  jar  should have one internal  label  and two  external  labels.
The  internal  label  should  be written with  indelible  ink on waterproof,
100-percent rag  paper.  Paper  with less than  100-percent rag  content or
that  is  not waterproof will fall  apart in the 70-percent  alcohol mixture,
and the  ink will dissolve.   The  two external  labels  should be  preprinted
 and should  be labeled with an indelible marking pen.   One  label  should
be attached  to the jar  and the second should be  attached  to  the  lid of
 the jar.  All three labels must include all information recorded on the
 field  data  tag, plus all other information needed to  ensure proper identifi-
 cation of the sample  (e.g., if more than one  jar, label  them 1 of 2, 2
 of 2,  etc).

      Keep  all  jars of a given  sample  together (if  more  than one), and  all
 replicate samples from a given station together.   As  the samples are shelved
 prior  to sorting,  each  should be cross-referenced  to the field log sheet.
 At this point the sample  custodian  should date and initial the rescreening
 section of the sample-track ing  form  for each station.   ^Store washed  samples
  in an upright position, at a cool temperature, and away  from direct sunlight.
 Storage  should be in  a secure place,  where sample containers are  not  exposed
  to breakage or tampering.
                                     224

-------
                                                                 Sediment
                                                                  Infauna
     Several techniques  can  be  used to sort organisms from sediment.   The
most common  technique involves placing  a small amount  of the sample  into
a glass  or  plastic petri dish and using a pair  of jewelers forceps to  sort
through the  sample  in a systematic  manner, removing  each organism.   This
entire  process  should be done while viewing the sample  through a 10-power
dissecting microscope.  Care must be  taken that  enough liquid is  present
in  the  pet  i  dish to completely  cover the sample,  otherwise, reflections
from the sediment/liquid interface  will  cause distortions and the  sorter
may miss some organisms.  £* h peu i  dish of material  should be sorted
twice to oe  surb that all  organ.sms are removed.

     A second  sorting technique is  a  flotation method, which is particularly
effective when the  sediment residue  is primarily coarse sediment grains
containing  small   amounts of organic  matter (e.g., wood fragments, leaf
debris, sewage sludge).   The  sample  is first washed with fresh  water in
a  large flat tray.  The  less dense material that becomes suspended in the
fresh water (organic material, arthropods, and most soft-bodied organisms)
is  carefully poured into a sieve, and is sorted using the  standard  technique
described above.   The remaining material is covered with liquid  and sorted
using a 5-power  self-illuminated hand lens.  Organisms remaining  in this
portion of the sample generally  include molluscs and some  tube-dwelling
or encrusting organisms  that are associated  with sand grains.   Because
 it is difficult to see extremely small organisms  with  the 5-power hand
 lens, the sorter  must remove all molluscs  and polychaete  tube  fragments
 for closer inspection.  All material  collected from this  portion  is placed
 into  a labeled  sample  jar and viewed under  a  10-power dissecting  microscope
 to remove organisms from  tubes and to ensure that the molluscs  were  alive
 when  captured.

      Both sorting techniques expose  the sorter to  alcohol  fumes.  Because
 these fumes can be irritating to some  people, the  sorting process can  be
 done using  fresh water.  However, as  each portion of  the  sample is sorted,
 it should be drained and  returned  to the alcohol  solution  immediately.
                                    225

-------
                                                                  Sediment
                                                                   Infauna
     Each  sample should be  sorted  by only  one  person.  Organisms should
be sorted Into the  following major taxonomic groups:   Annelida, Arthropoda,
Mollusca,  Echinodermata, and miscellaneous phyla  (combined).  All organisms
should be placed in large  vials containing 70-percent  alcohol  solution.
The  exception  is  Ophiuroidea, which require air-drying  for  identification.
Removal of the majority of arms from certain Ophiuroidea  (e.g., Amphiuridae)
permits easier identification.  This procedure may be  performed by experienced
sorters torn nmize identification time.  (Note:   Special  handling of Ophiuroi-
dea  should be  conducted  after biomass  analyses, if biomass analyses ure
performed.)  Each vial containing a  major taxonomic  group should  have an
internal  label  listing the survey name, station designation, water depth,
date sampled, and field screen size.   All  vials in the  same sample  should
be  stored  in a  common container and  immersed  in the  70-percent alcohol
solution.  To reduce evaporation of  alcohol, vial and  container  lids can
be sealed with plastic tape.

     When  required, biomass estimates for the major taxonomic groups should
be made prior to identifying the organisms to the  species level.   However,
it is  recommended that taxonomists examine the major taxonomic groups before
biomass measurements are made, to ensure that sorters  have correctly grouped
all  individuals and fragments.  Biomass should be estimated  to the nearest
0.1  g  (wet weight).  All specimens of taxa within  the  following major groups
should be composited for biomass analyses:  Annelida  (principally polychaete
worms), Mollusca (principally bivalves, gastropods and aplacophorans),
Arthropoda (principally crustaceans),  Echinodermata  (principally asteroids,
ophiuroids, echinoids, and holothuroids), and miscellaneous  taxa (combined).
These  five categories generally are adequate to characterize  the  standing
stocks of the major infaunal groups.  They also are  sufficiently  distinct
from each other to permit  proper  assignment of fragments to each of the
groups.  All fragments should be placed in their respective  major  taxonomic
groups prior to weighing.
                                     226

-------
                                                                  Sed iment
                                                                   Infauna

     There are  several  major problems  associated with the  collection  end
interpretation of biomass information.  Some taxa lose weight  when  immersed
in preservative  fluids, while  others gain weight (Howniller 1972; Lappalainen
and Kangas 1975; Wiederholm and  Eriksson  1977; Mills et  al.  1982).   For
this  reason, the most accurate  biomass estimates are performed  on live
material.   However,  it is rarely  practical to  sort and weigh live  specimens.
Accurate  measurements of .iom*ss  are further compromised  by evaporation
from the specimens while they  are on the balance.

     Several methods  of measuring jiomass are possible.  One technique
is to estimate the difference  in weight  of a  tared beaker filled with preser-
vative  before and after organisms are  placed  in the beaker.  The  individual
organisms are not blotted prior to weighing, and as  few individuals as
possible are transferred to the weighing  container.  These procedures minimize
the transfer of  fluids held within a pile of individuals.   This  technique
can  be  used for preserved or live animals, and  appears to  introduce the
least amount of  variation into the weighing  process.

     A  second technique for  biomass  determination consists of air-drying
the organisms on  absorbent pa'per for  a  specific length of time (e.g., 5 min).
Because 70-percent ethanol  is  volatile,  small  variations in drying time
may increase the  errors associated with  the  weight measurements.  A container
open  at one end and covered at the  other end with a 0.25-mm mesh screen
(maximum mesh opening)  can  be used to  hold the organisms  for weighing.
After the tare weight of the container  is measured,  the  animals are carefully
placed  into the  container.   The container with organisms is then placed
on a  paper towel and allowed  to air  dry for exactly  5 min  prior to weighing.
The weight of the organisms is obtained  by subtracting the weight of the
container with the organisms  from the  tare weight of the container.   Extremely
 large organisms (e.g.,  large  molluscs  or asteroids)  should be  weighed indi-
 vidually.
                                     227

-------
                                                                  Sediment
                                                                   Infauna
     After  biomass  estimates are completed,  idertification and  counting
of the organisms may begin.  Identifications  shojld be to the lowest taxonomic
level possible,  usually the species  level.   For  incomplete specimens,  identify
all fragments, but enumerate only the  anterior ends  (i.e.,  heads).  All
identifications require the use of  binocular  dissecting and compound micro-
scopes.  If possible,  at  least two pieces of literature  should be  used
for each species identification, one of  which  should be tne original  descrip-
tion.  Moreover,  each  species identification should be  checked  \gainst
a  reference specimen from a verified  voucher museum  c"'lection.

     After completing an identification,  all organisms of a single species
should be  placed In  a vial containing 70-percent alcohol.  All vials  should
be stored  in a common jar with the original  internal label (from the field),
and  immersed  1n 70-percent alcohol.   Each vial  should  contain an internal
label  with the  following  information:  species name, number of specimens,
survey name,  station number, replicate number, collection gear, water depth,
date of collection, latitude, longitude, and initials of the taxonomist.
Any specimens removed fron the sample jar and placed in  the reference collection
should  be  so  noted  (species, number) on the sample  identification sheet.

      Each taxonomist  should  record initial  identifications and counts in
a bound, hardcover  notebook, which should also  include notes and comments
on the  organisms in each sample.  Upon completion of the  sample, the data
should be transferred in ink to the sample data sheets and double-checked.
The taxonomist  should then sign and date the sample data  sheet.  All notebooks
 should be kept  in the laboratory at all times so the laboratory  supervisor
can check questionable identifications and  follow the progress  of each
 sample.   Erasures from notebooks or data sheets are not permitted.   Simply
draw  a  horizontal  line through incorrect  data,  initial  and  date  it, and
 indicate why the data were removed.

      Calibration and  Preventive Maintenance—The analytical  balance used
 for biomass determinations should  be calibrated weekly, at  a minimum.
                                     228

-------
                                                                   Sed iment
                                                                    Infauna

The  balance and all microscopes should  be serviced at rec.ular  intervals.
Annual  service and inspection is adequate in most cases,  unless the manufacturer
recommends otherwise.

     Taxonomic  identifications should be consistent within a given  laboratory,
and with the identifications of other workers.   To that end, at least three
individuals of each taxon  should be  sent for  v :r ification to  recognized
experts in museums  and/or academic institutions.   The verified  soecimens
should then be  placed in the permanent reference museum.  Continued co1  •ction
of a verified species does  not require additional  e.-oert  .erif icacion.
Participation of the laboratory staff in a regional taxonomic  standardization
program is required (if available) to ensure regional consistency and accuracy
of identifications.  The reference collection should  be started and completed
as early as possible during the project.

     All  specimens in  the reference collection should be in labeled  vials
that are segregated by species and sample.   For example, there may be three
labeled  vials  of Gemma gemma, one  from  each of three samples.   More  than
one specimen may be in each vial.   The labels placed  in these vials should
be the same as  those used for specimens in the sample jars.  It is important
to complete these labels because future workers  may not  be  familiar with
the  survey, station locations,  and other details of the work in progress.
In addition, the reverse side of the label  should contain information about
the confirmation of the identification by experts in  museums or other insti-
tutions (if appropriate).  Such Information would Include the outside consul-
tant's name, institution, and date of verification.  All vials for a  given
species should  be placed in a single  jar filled  with  alcohol.  To redur
evaporation of  alcohol, the lids of vials and jars can be sealed with pi a'
tape.   The species (or other taxonomic designation) should be clearly wr
on the outside and on a  large  internal label.  Reference specimen*
be archived alphabetically within major taxonomic groups.
                                     229

-------
                                                                 Sediment
                                                                  Infauna

     A computer  listing of each  species name, the  name and affiliation
of the person who  verified the identification, the location of the  individual
specimen  in the museum,  the  status of  the  sample if  it  has been  loaned
to outside experts, and references to pertinent  literature should be  maintained
by  the  laboratory performing identifications for the monitoring  program.
This listing should be available near the  collection.

     Reference specimens are  invaluable, and should be retained  at  the
location where the identifications were  perfo,  ed, in the  offices of  the
funding agencies, or  at a museum  with  long-term storage capabilities.
In  no instance should this portion  of  the collection be destroyed.   One
person  should be  identified  as the curator of the  museum collection  and
should be responsible for its  integrity.   Its upkeep will require periodic
checking to  ensure  that alcoho> levels are adequate.   When refilling  the
jars, it is  advisable to use full-strength alcohol  (i.e., 95 percent),
because the  alcohol  in the 70-percent solution will  tend to evaporate more
rapidly than  the water.

      Quality Control Checks —It is recommended  that  at  least 10 percent
of each  sample be re-sorted  for QA/QC purposes.  Re-sorting is  the  exami-
nation of  a  sample or subsample  that has been sorted  once and  is considered
free of  organisms.  Each sample  should be re-sorted  by  someone  other than
the original sorter.  The 10-percent  aliquot should  be  taken after the
entire  sample has been spread  out in an  enamel pan.   It is critical that
 the aliquot be a representative subsample of the total sample.

      A  partial re-sorting of every sample should ensure  that all gross sorting
 errors  are detected.  In addition, it  should give  added  incentive to  sorters
 to  process every sample accurately.  An  alternate 10-percent QA/QC approach
 of re-sorting one entire  sample out of every  10  processed does not have these
 advantages.  Because 9  out of every  10  samples  are not re-sorted, the proba-
 bility of randan gross  errors going  undetected  increases, and the incentive
 of  sorters to process every sample accurately decreases.
                                      230

-------
                                                                Sed imen t
                                                                  Infauna
     In addition to efficient  sample sorting, consistent identifications
of organisms among individuals and among  sampling programs  are critical
to the collection  of  high quality data.   Consistent identifications are
achieved  by  implementing  the procedures discussed below  and  by maintaining
informal, but  constant, interaction among  the taxonomists working  on  each
major group.  One important procedure  is to verify identifications by camp rison
with the  reference collection.  To ensure that identifications are  correct
and consister.    5 percent of all samples identified  by one taxonomist should
be re-identicied by another taxonomist  who is also qualified to identi'•/
organisms in that major  taxonomic  group.   Differences in  opinion should
be recorded  and given to the senior  taxonomist.  It  is  the duty of the
senior taxonomist to decide upon the  proper  identifications.  The senior
taxonomist  may also decide whether  the  taxonomic level  to which  a given
organism is  identified is appropriate.  If it is not, the senior taxonomist
may decide  to drop back to a higher taxonomic level,  or to  further refine
the taxonomy of that group through  additional study.

     Three  kinds of material  should be  archived  during the  course of any
sampling program:   sediment residue (after sorting), identified organisms,
and  reference  specimens.  First, the  sediment residue  (i.e., "grunge")
should be characterized, and  a representative  aliquot  should  be  archived.
The  characterization should  include a description  of the  major  sediment
component (e.g., coarse to  fine  sand, wood,  shells, or  polychaete  tube
fragments, organic material,  sewage sludge) and the volume of the total
amount of material.  An 8-dram vial filled  three-quarters and  topped off
with 70-percent alcohol  is  sufficient for  archiving these aliquots.  Each
vial  should be labeled  with  the   same information found  on  the internal
sample label.   All vials should be-tightly  stoppered and stored  together
 in a container  filled with  70-percent  alcohol.   Black electrical tape wrapped
 tightly around the lid of  the large container  will improve the  seal  and
 insure that the alcohol  does not evaporate during storage.   These archived
                                     231

-------
                                                                 Sediment
                                                                  Infauna

ali'qucts  of sediment residue  provide  a qualitative reference  sample should
questions concerning sediment characteristics arise later.

     When  all  identification  and QA/QC procedures are completed, the jars
containing the vials of identified  species should be topped off with 5-percent
glycerine/7n-percent alcohol.   The  lids should then be sealed tightly with
black electrical  tape to prevent  evaporation.   All  sample iars  should be
placed  in containers filled with 70-percent alcohol for long-term storage.
The containers should be fitt  1 with  a  tightly  sealed lid, and electrical
tape  should again be used to seal  the joints.   Each container should be
labeled clearly with the survey name,' date,  and number and type of  samples
within it.

     Corrective Action—Foil owing QA/QC procedures discussed  earlier, each
sample residue should be checked  for complete  or nearly complete  removal
of  organisms.  Thus, each sample elicits  a decision concerning a possible
re-sort.   When a sample is found  that does  not  meet the specified  removal
criterion, it should be re-sorted.  In addition, the previous  three  samples
processed  by that sorter should be re-sorted (regardless of their  aliquot
re-sort  results)  to ensure  that sorting  efficiency  has been adequate.
If  one of  the three samples is  incompletely  sorted, all samples processed
by  that  individual  should be re-sorted.

     When a taxonoaic  error  is  found, it  is  necessary to trace all  of the
work  of  the taxonomist  responsible for  the error, so as to  identify those
samples  into  which the  specific error may have  been  introduced.  This  process
can be very time-consuming.   However, upon completion  of   all taxonomic
work,  few (if any) taxonomic errors should remain in the data  set.  Avoiding
 errors through the constant  interchange of  information  and  ideas among
 taxonomists is the best way to minimize lost time due  to misidentification.

      Data  Quality  and Reporting—Generally, a sample sorting efficiency
 of 95 percent is considered acceptable.  That  is,  no more than 5  percent
                                     232

-------
                                                                 Sediment
                                                                  Infauna
of the  organisms in a given  samp'e  are missed by the sorter.   Similarly,
specimen identifications can reasonably be expected to be at  least  95 percent
accurate.   All  organisms must be  identified to the lowest possible taxon,
and to species level whenever  possible.  In  cases where  the  identity of
a species  is  uncertain, a species number will suffice (e.g.,  Macana sp.l,
Macoma sp.2).  Numerical designations must be  consistent  throughout  the
study.  Data  for each replicate  sample should  be reported  as  numbers of
individuals/m2 for each species and  as biomass (nearest 1.1  g wet  weight/m2)
for each major taxonomic group.
                                     233

-------
3IOACCUMULATION/TRAWL ANALYSES

     QA/QC  procedures are presented  in  this section for  the  following  tnree
tissue/trawl variables:-

     •    Priority pollutant metals

     •    Priority pollutant organic  compounds

     •    Demersal fishes and epibenthic macroinvertebrates.

     Or;  nlsho  to be  analyzed for these variables generally will  be collected
jsing  an  otter  trawl.  Operation of  an otter  trawl is discussed e-rlie'-
in  the  general  methods section.
                                       234

-------
                                                          Bioaccumulation
                                                 Priority Pollutant Metals
Priority Pollutant Metals
Field Procedures--
     Col1ection--The major difficulty  in  trace metal analyses of tissue
samples is controlling contamination of the  sample after collection.   In
the  field,  sources of contamination  include sampling gear, grease from
winches or cables, engine exhaust,  dust, or  ice used  for cooling.   Care
mijst  be  taken  during handling to avoid these and any otier  possible sources
of contamination. For example, during sampling the sli > should  be positioned
such  that the  engine exhausts do not  fall on deck.  To avoid  contamination
from melting ice, the samples should be wrapped in aluminum  foil  and  placed
in watertight  plastic bags.   The  outer skin of  the fish or shell of the
shellfish acts  as protection against metals contamination from  the  aluminum
foil.

     Sample resection and any subsampling  of the organisms  should be carried
out  in a controlled environment  (e.g., dust-free room).  In most  cases,
this  requires  that the organisms  be  transported on ice  to  a laboratory,
rather than being resected on board  the sampling vessel.  It is recommended
that whole organisms not be frozen  prior  to  resection if  analyses will
be conducted only on selected tissues,  because freezing may cause  internal
organs to rupture and contaminate other tissue.   If organisms are eviscerated
on board the survey vessel, the  remaining  tissue (e.g., muscle)-may be
wrapped as described above and  frozen.

      Resection is best performed under "clean room" conditions. The  "clean
room"  should have positive  pressure and  filtered air.  The  "clean room"
should also be entirely metal-free and  isolated  from all samples  high  in
contaminants (e.g., hazardous waste).   At a minimum,  care  should  be taken
to  avoid contamination  from  dust, instruments, and  all materials that may
contact the samples.  The best  equipment to use  for  trace metal  analyses
 is  made of quartz, TFE, polypropylene,  or polyethylene.  Stainless steel
that is resistant to corrosion may be used if necessary.  Corrosion-resistant
                                    235

-------
                                                          Bioaccumulatlon
                                                 Priority Pollutant Metals

stainless  steel is not magnetic, and thus can be distinguished from othe^
stainless steels with a magnet.   Stainless steel  scalpels have been  found
not  to  contaminate mussel samples (Stephenson 1979).  However, low concen-
trations of heavy metals in other biological  tissues (e.g., fish muscle)
may  be  contaminated significantly  by any exposure to stainless  steel.
Quartz utensils are  ideal  but expensive.  To control contamination  when
resecting  tissue, separate  sets of utensils  should be used  for removing
outer tissue and for removing tissue  for  analysis.  For bench liners and
bottles, borosilicate glass would be  preferred over plastic  if tr-ce organic
analyses are to be  performed on the same  sample.

     Resection should be conducted by or  under the  supervision of a competent
biologist.  For fish samples, special care must be  taken to avoid contaminating
target  tissues  (especially muscle) with slime and/or adhering sediment
from the fish exterior  (skin) during resection.  The  incision  "troughs"
are  subject  to  such contamination.  Thus,  they  should not be included in
the  sample.   In the case of muscle,  a "core" of  tissue is taken  from within
the  area boarded  by the  incision troughs, without contacting  them.   Unless
specifically  sought as a sample, the dark  muscle  tissue  that  may exist
in the vicinity of the  lateral line should not be mixed with  the  light
muscle  tissue that consitutes the rest  of the  muscle  tissue mass.

      Prior to  use, utensils  and bottles should be  thoroughly cleaned with
a detergent solution, rinsed with tap water,  soaked in acid, and then rinsed
with metal-free water.  For quartz, TFE, or glass containers,  use 1+1 HNQ-3,
1+1  HC1,  or  aqua  regla (3 parts cone HC1  + 1  part  cone HNQ-3) for soaking.
For plastic  material,  use 1+1 HNQ-3  or 1+1 HC1.   Reliable soaking  conditions
are 24  h  at  700 C (APHA 1985).  Do not use chromic acid for cleaning any
materials.  Acids used  should  be at  least reagent grade.  For  metal  parts,
 clean  as  stated for glass  or  plastic, except omit the acid soak  step.
 If trace organic analyses are to be  performed  on  the same samples,  final
 rinsing with  methylene chloride  is acceptable.
                                     236

-------
                                                           Bioaccumulation
                                                 Priority Pollutant  Metals

     Sample  size  requirements can  vary with tissue  type (e.g.,  liver or
muscle) and detection  limit  requirements.   In  general,  a minimum sample
size of 6 g (wet weight) is required  for  the  analysis of all priority  pollutant
metals.  To allow  for duplicates,  spikes,  and  required reanalysis, a sample
size  of 50 g (wet weight) is recommended.   Samples can be stored in  glass,
TFE, or high-strength polyethylene jars.

     Processing—Samples should be frozen after resection  and kept at -200 C.
Although specific  holding times have  not been recommended by U.S. tPA   a
maximum holding  time  of  6  mo (except  for  mercury samples,  which  sho  Id
be held a maximum of 28 days) would  be  consistent with that  for water  samples.

     When  a  sample is thawed, the  associated  liquid should be maintained
as a part  of  the  sample.   This  liquid  will contain  lipid  material.   To
arfoid  loss of moisture  from the sample, partially thawed samples  should
be homogenized.   Homogenizers used to grind the  tissue should have  tantalum
or titanium parts  rather than stainless steel  parts.  Stainless steel  blades
used during homogenization have  been  found  to be a source  of nickel  and
chromium contamination.

Laboratory Procedures--

     Analytical  Procedures—Priority pollutant metals  should be analyzed
according  to the procedures described in Tetra Tech  (1986b).  Digest  the
homogenized  sample prior to analysis,  using the acids specified  in the
procedure.  The digestate  can then be analyzed by flame  Atomic Absorption
Spectrophotometry (AAS), graphite furnace AAS, or Inductively Coupled Plasma
 (ICP), depending on the sample concentration  and required  detection limits.
 ICP  can be used to screen  samples for elements that  are present in  relatively
 high concentrations.   Most elements  will require more sensitive  analysis
 by graphite  furnace  AAS.   Mercury analysis must be performed  by cold vapor
AAS.
                                     237

-------
                                                           Bioaccumulation
                                                 Priority Pollutant Metals

     Calibration  and  Preventive Maintenance—Because  contamination can
be the limiting factor for reliable  quantitation of trace metals in tissue,
several field and  method blanks should  be  analyzed to establish contamination
variability.  At least three blanks, or a  blank every 20  samples, should
be collected at each step  of sample  preparation to evaluate contamination
variability.  The  minimum blanks to  analyze  are:

     a    Bottle  blanks    Bottles should  be opened and  rinsed with
          concentrated HN03.   The HN03  rin-,e should be digested and
          analyzed with each batch of samples

     0    Resection blanks  -  After routinely cleaning  the utensils
          used for dissecting  and  before resecting  the next  sample,
          the  utensils that have  contacted  the  tissue to be analyzed
          should be  rinsed with metal-free water.   A rinsate  should
          be  collected and the volume  recorded.  This rinsate blank
          should be  analyzed at least once and preferably  one blank
          with  each batch of samples.   The  total micrograms  of each
          element in the  rinsate should be recorded.   This blank will
          allow evaluation of  the cleaning technique and provide correction
          values If  necessary.

      •   Digestion  (or  method) blanks  - Blanks containing  the  same
          quantity of acid as each sample should be digested  and  analyzed
          with each batch of samples.

 Preliminary blanks and  samples  should  be analyzed prior to resecting and
 analyzing the tissue samples to ensure that  the  blank concentrations and
 variability are low enough to  establish adequate assessment of tissue concen-
 trations.  Blanks should  continue  to be analyzed with each batch of samples
 to monitor  sporadic contamination.

      All analytical  instruments  must  be calibrated daily and  each time
 the instrument is set up.  Calibration procedures should  follow those for
                                      238

-------
                                                           Bioaccumulation
                                                 Priority Pollutant Metals

t^e  specified method for each  analysis.  Calibration standards must be
prepared using the same concentrations of  acids as those which  will  result
in the samples following sample preparations.

     After an  instrument has been  calibrated, the accuracy of the  initial
calibration should be verified by the  analysis of certified control  solutions
at a  frequency of 5 percent or every 2 h during an analysis run,  whichever
is more  frequent, and after running the  last  analytical sample.  If a  certif .ed
control solution is not available, a standard  should  oe  used.  Analyte
for the standard ;hould be from a different  source than  that  used in the
standards for  tlie initial  calibration.   If the deviation of the continuing
calibration verification  is  greater than the calibration control limits
specified in the method, the instrument must be recalibrated, and the preceding
10 samples reanalyzed.

     All  equipment  should have  scheduled  routine preventive maintenance,
and a record of all maintenance performed  should be  noted in a logbook.
Critical  spare parts  should be kept on  hand.

     Quality Control  Checks—Analyze appropriate standard  reference materials
 (SRM)  if  available  (e.g., U.S. EPA Trace Metals  in  Fish  Tissue, NBS Oyster
Tissue, or,  if analyzing  fish liver,  NBS  Bovine  Liver)  to provide a check
on contamination, digestion efficiency,  and overall  accuracy of the analysis.
A minimum of  one SRM  should be analyzed  for each survey or 2 percent of
the  total number of samples (i.e.,  1 per 50 samples), whichever  is  more
 frequent).

      To estimate precision,  a  minimum  of 5  percent  of  the  total number
 of samples  should be analyzed  in duplicate or one  for each survey, whichever
 is more  frequent.   When more than 20 samples are to be analyzed for one
 survey, the  project manager may  choose to implement a  program of triplicate
 analyses.  The overall  percentage of replicates should  be at least 5 percent.
 Analyze samples spiked before digestion at the same frequency as duplicates
 to estimate recovery.  Add  spike  concentrations approximately  equal  to
                                     239

-------
                                                           Bloaccumulation
                                                 Priority Pollutant  Metals

tne  concentration found  in  the unspiked  sample.   An acceptable range  in
spine concentrations  is 0.5 to 5 times the sample concentrations.  Limited
sample size may necessitate reduction in the number of QC samples for  bioac-
cumulation monitoring.

     The blanks listed  under Calibration Procedures and Preventive Maintenance
should be carried through  all the appropriate steps.   Ideally, the concen-
trations of the  bottle  blanks, dissection blanks, ancj  digestion  blanks
are  less than the detectir • lim-t and no correction  of  sample results  is
performed.  If  there  is  extensive  contamination of blanks (>30 percent
of sample value)  and if there is enough  sample, the batch of samples associated
with the blank  should be reanalyzed.  The data should be corrected by the
data users  for the blank  values  between the required  detection  limit  and
the  control limit.

     For  ICP analysis, additional QC  checks should  include  an  interference
check  sample  to verify interelement and  background correction  factors.
For graphite  furnace AAS, additional QC checks should  include duplicate
 Injections with  the mean value  reported.  The  relative standard  deviation
of  the readings should  be within  control limits,  or the sample should be
rerun.   If the  predigestion  spike recovery is  poor, the  digestate  should
be  spiked and results compared to  the  predigestion spike results.   If the
digestate spike recovery  is as  poor  as  the  predigestion recovery,  the method
 of standard additions should be used  to determine digestate concentrations.

      Corrective Action—If the concentration of the field or method  blank
 is  greater than the  required detection  limit,  then all steps in the  sample
 handling should be reviewed.   Many trace metal contamination problems  are
 due to airborne dust.  Thorough cleaning  of all utensils between processing
 each  sample is important to minimize cross-contamination.   High zinc blanks
 may be due to airborne dust or galvanized iron, while  high chromium  and
 nickel  blanks  indicate  contamination from stainless steel.   In the field,
 the use of mercury-filled thermometers should be avoided.  Broken thermometers
 can be a  potential  source of severe mercury contamination.   In the laboratory,
                                     240

-------
                                                           Bioaccumulation
                                                 Priority  Pollutant Metals

samples for mercury analysis should  be  isolated from items such as polarographs
or COO reagents.

     Poor  replication  may be caused  by  inadequate mixing of the sample
before taking  aliquots, inconsistent contamination, inconsistent digestion
procedures, or instrumentation problems.

     Poor  performance  on the analysis  of the Standard Reference  Material
(SRM) or poor  spike recovery  may be caused  for  the same  reasons  as poor
duplication.   However,  if duplica. >  results are acceptable, then  poor  SRM
performance or poor spike recovery may  be  caused by loss of analyte during
digestion  or  sample matrix  interferences  during analysis.  To check  for
analyte loss during digestion or   for low recovery due to interferences
during  analysis, spike the sample  after  digestion and compare the  analysis
to the predigestion spike.  If the results are different, then the digestion
technique  should be adjusted.   If the  results  are the same, then dilute
the sample  by  at  least a factor of five and reanalyze.   If spike recovery
is still poor, then the  method of standard additions, a matrix modifier,
or another  method is required.

     Sediment on the  outside of shellfish may contaminate the tissue to
be analyzed.  Proper cleaning of the  outside  of the animal  is critical.

     Data  Quality and  Reporting—Measurements should be reported  as  ug/kg
to a maximum of three significant figures  on  a wet-weight basis.  Detection
limits can  vary widely because of methods  and Instrumentation.   The analytical
method should be consulted  to determine  expected detection limits, precision,
and  accuracy. Further guidance on  analytical detection limits can  be found
Tetra Tech   (1985b).  '

     The data report  should  include  duplicate,  spike, and blank  results.
Data should   not be blank-corrected.   In  addition,  the  laboratory data summary
should  include:

                                    241

-------
                                                             Bioaccumulation
                                                   Priority  Pollutant Metals


     •    Digestion procedures


     0    Quantity of sample digested  and  final  dilution  volume


     •    Instrument detection  limit  for each element


     t    Blank associated  with sample


     •    Deviation from the prescribed methods


     •    Problems associated with  analysis.


For  a  more thorough QA review,  additional  documentation (e.g.,  calibration
curves) may be requested.
                                      242

-------
                                                          Bloaccumulation
                                      Priority Pollutant Organic Compounds

Priority Pollutant Organic  Compounds

Field Procedures--

     Col1ection--In  the  field, sources  of  contamination  include sampling
gear, grease  from winches or  cables, engine  exhaust,  dust, and  ice  used
for  cooling.  Efforts should  be made  to minimize  handling and to avoid
sources of contamination.   This  will  usually  require resection of  tissue
to be performed  in a controlled  environment (e.g., laboratory).  For examp'.c
to avoid contamination from ice, the  samplps should be w jpped  in aluminimum
foil,  placed in watertight  plastic bags and immediately  iced.  Organisms
should not be frozen prior to resection if analyses will be conducted  only
on  selected  tissues because freezing may cause internal organs to rupture
and  contaminate  other tissue.   If organisms  are  eviscerated on board the
survey vessel, the remaining tissue  (e.g., muscle) may be-wrapped as described
above and frozen.  Limited sample size can be a  problem when  using  small
organisms, but  an attempt should be  made  to collect at least  30 g  of tissue
for  each organic priority  pollutant analysis.  Additional  sample  (30 g)
is needed for each spike, duplicate,  reanalysis, and archive sample.

     Processing—To  avoid cross-contamination, all equipment used  in sample
handling should be thoroughly  cleaned before each sample  is processed.
All  instruments must be  of  a  material  that can be easily cleaned (e.g.,
stainless steel, anodized  aluminum, or borosilicate  glass).   Before the
next sample  is processed,  instruments  should  be washed  with a  detergent
solution, rinsed with tap water, soaked  in high-purity  methylene chloride,
and finally  rinsed  with  reagent water.  Work  surfaces should be  cleaned
with 95-percent ethanol, and  allowed to dry  completely. If metals are  to
be analyzed, instruments may  be washed with acid  prior to the methylene-
chloride rinse.   Each fish, crab, and mollusc should  be handled with  clean
 stainless  steel  or  quartz  instruments  (except  for  external surfaces).
The specimens should  come  into  contact with precleaned  glass surfaces  only.
 Polypropylene and polyethylene  surfaces are a potential  source of contamination
 and should not  be used.   To control contamination  when dissecting tissue,
                                     243

-------
                                                          Bioaccumulation
                                      Priority Pollutant Organic Compounds

separate sets of utensils should  be  used for removing  outer tissue  and
for dissecting tissue for analysis.   For fish  samples, special care must
be  taken to avoid contaminating target tissues  (especially muscle) with
si ire and/or adhering sediment  from the fish exterior (skin) during resection.
The  incision "troughs" are subject to such contamination.  Thus, they should
not be included in the sample.   In the case of  muscle, a "core" of  tissue
is  taken from within the area boarded by thp  incision  troughs, without
contacting them.   Unless specifically sought as a  sample,  the dark  muscle
tissue that may exist  in the  vicinity of the  lateral l1050 C may be substituted for
the solvent rinse if only volatile organics are to  be  analyzed.

      The U.S. EPA and  other  federal  agencies  (e.g., National  Bureau of
 Standards)  have  not jet  provided specific guidance  regarding  holding times
 and temperatures for tissue samples  to be analyzed  for semi-volatile  organic
 compounds.  Until U.S.  EPA develops  definitive guidance,  the  following
 holding conditions should  be observed.   Resected  tissue  samples should
 be maintained at -200 C and extracted as soon as  possible, but within  10
 days of sample receipt.   Complete analyses should be performed  within 40
 days.

      When  possible, the  entire  sample should be used for analysis.  There
  are  several acceptable methods of homogenization of a thawed tissue  sample.
  If a micro-grinder (e.g.,  a  Tissuemizer)  is  used, care must be  taken to
                                     244

-------
                                                           Bioaccumulation
                                      Priority Pollutant Organic Compounds

ensure that  it is thoroughly  cleaned between each use.   This  usually entails
disassembly  of the unit.   A scapel can be used to cut the  tissue  into  small
pieces.  The method chosen will depend upon sample size, time, and resources
available.   Devices with  large  surface areas (i.e., blenders, meat grinders)
should  be avoided,  as they are difficult to  clean  and a  small sample is
difficult to remove after grinding.

Laboratory Procedures--

     Analytical Procedures--Priority pollutant organic compounds  shouU
be analyzed  according to  the  procedures described in Tetra Tech  (1986b).
Preparation of  the  tissue will depend upon sample size,  extraction method,
and sample type  (fish muscle,  whole organism,  or composite).  Avoid use
of plastic  or  rubber equipment that may come in contact with  the sample,
as these can be  a source  of contamination.

     A modified U.S.  EPA Method 624  (Hiatt 1981) produces adequate spike
recoveries  and detection  limits for volatile  compounds.  Analysis  of the
semi-volatile  organic compounds  involves a solvent extraction and  one or
more cleanup procedures.

     A Soxhlet  extraction  with at least  a gel  permeation column cleanup
provides efficient extraction  of organic compounds  of interest and eliminates
many interfering coextracted  materials.

     Cross-contamination should be  avoided during all steps  of  analysis.
All glassware  should be clean.   Injection  micro-syringes must  be cleaned
well  between  uses.  If separate  syringes are  used  for  injection of standard
solutions, bias  between syringes  should be  accounted.   Carryovers can  occur
when  high-  and low-level samples are analyzed  sequentially.   Analysis  of
an appropriate  solvent blank following a  high-level  sample may be necessary
to check for carryover.
                                     245

-------
                                                           Bioaccumulation
                                       Priority Pollutant Organic  Compounds

     Recommended  detection limits  are based on a minimum 20-g wet-weight
sample size with an additional  10 g  (wet  weight)  required  for  a  separate
analysis of volatile compounds.  For  most  of  the volatile organic compounds,
detection limits between 5 and 10 ug/kg wet  weight are recommended.  Detection
limits  of 10 ug/kg (wet weight)  are recommended for aromatic hydrocarbons
and phthalates analyzed  by  GC/MS.    Detection limits of  10-20  ug/kg (wet
weight)  are recommended for chlorinated hydrocarbons and halogenated  ethers
analyzed by GC/MS.  Detection limits of 50 ug/kg  (wet weight) are recommended
for pesticides analyzed by GC/MS.  A detection limit of 20 ug/kg (wet weight)
is recommended for PCBs analyzed by  GC/Et .  If pesticides  are  analyzed
by GC/EC, a detection  limit  in the  range  of 0.1-5 ug/kg (wet weight)  is
recoimended (Tetra Tech 1985b).

     Actual attainable detection limits will  vary with sample size, cleanup,
extraction method, final volume, and amount and  nature of any co-extractive
compounds.

     Calibration  and Preventive  Maintenance—Before beginning analysis
of samples, a calibration curve bracketing the working range must  be estab-
lished.  This calibration should  be repeated after all  major  equipment
disruptions.  Calibration checks of  the GC/MS system  at the beginning  of
the day and at least  every  12  h will establish that  the  instrument's response
is  in control.  Specific tuning  criteria (e.g., OFTPP,  BFB)  are given  in
each method.  Calibration checks  of the GC/ECO system  should  be done at
the  beginning of the  day and verified at least every  6  h. These are only
recommended minimum frequencies  and  the nature of the  samples may  necessitate
more  frequent verification.

      A  routine QC check  for each lot of analytical  reagents  used in extraction
can  monitor a potentially  serious  source  of  contamination.  Also,  each
 lot of alumina,  silica gel, sodium sulfate, or  Florasil used  in extract
 cleanup should be tested and cleaned as necessary.   Any resins or  charcoals
 used should  also be monitored.   Surrogate mixtures  have also  been found
 to contain impurities and should be  verified  prior to use.  The fatty  acid
                                      246

-------
                                                          Bioaccumulatlon
                                      Priority Pollutant Organic Compounds

content of many tissue samples may overload the cleanup  columns.  The columns
should be calibrated and monitored  to enable  the  compounds of  interest
to be consistently collected  in the proper fraction.

     Equipment should  be maintained and  serviced  routinely according to
manufacturer's  instructions and good laboratory practices.  Logbook  records
should be kept  to document maintenance for each measurement device.

     Total Extract.^ble Lipids--The amount  of tissue extract required to
determine total ext  actable  lipid content will depend upon the sensitivity
of the  balance used and the amount  of lipid  in the  organism.  A maximum
of 1/200 of the original extract from 20 g of  tissue should be sufficient
for  this determination.  Ideally, using an electrobalance able to measure
to 0.1 ug, 1/10,000 of a 20-g extract is more  than  sufficient.   A volume
should  be used so as to measure  approximately 50  times the  smallest unit
       i
measured by the balance.  If a large percentage of the extract  is  used
for  lipid determination (e.g., >1/200), the  results of  further  analyses
on the remaining extract should be  adjusted  accordingly for  this  loss of
material.

     Quality Control  Checks—To  monitor precision  and accuracy, duplicate
and matrix spike analyses should be performed.   A  frequency  of  5 percent
of each or one  each  per survey, whichever  is more frequent, is the recommended
minimum.  When more than 20 samples are to be analyzed for one survey,
the  project manager may choose to implement a program of triplicate  analyses.
The  overall percentage of replicates  should be at least 5 percent.   Method
interferences can be  caused by contaminated glassware, reagents,  solvents,
or processing hardware.  These materials  can  be  monitored for  contamination
by processing of 5-percent method  blanks,  or one per batch, whichever  is
more frequent.   Addition of known  amounts  of surrogate compounds to each
sample before purging  or extraction will  serve to monitor preparation  and
analysis of samples.  An empty  tissue jar  processed,  handled,  and stored
as a sample should  serve as a  field  blank.   As  another measure of accuracy,
a standard reference material  should  be analyzed  if  available.  One  for
                                    247

-------
                                                           Bioaccumulation
                                       Priority  Pollutant Organic Compounds

each  survey or 2 percent of the total  number  of samples, whichever is more
frequent is the recommended minimum.

     Corrective Action—When results  of QC samples fall  outside of established
limits, several  courses of action  are available.   Contamination  in  the
lab  reagent blank  is  cause  for all positive  findings  of the  same compound
in associated samples  to be suspect.  If contamination  is extensive, reanalysis
of the whole associated  group of samples may be in order.  Blank contamination
should be kept to less  ' nan 10 percent of the sample values and, preferably,
below th2 detection  IT nit.   Contamination found in the field blank should
also be considered when  looking at associated sample data.  Extensive contam-
ination of a  lab or  field blank (>30 percent of sample value) should lead
to a detailed review of laboratory, sampling,  transport,  and  storage pro-
cedures.  Phthalates,  methylene chloride, and  toluene are common laboratory
contaminants  that may  be detected  in  blanks above the  method detection
1imit.

      Poor duplication  may be caused  by  inadequate mixing  of  the  sample
before removing aliquots, inconsistent contamination, inconsistent extraction
procedures,  or  instrument problems.   Further  replication  of an analysis
may be necessary to determine the reason for the poor results.

      Poor  spike recovery may be  caused  for the  same reasons as poor duplication,
or by matrix effects produced by  co-extracted  materials.   If the  spiked
compound  is added at a concentration much  less than that found in the sample,
recovery  may  be difficult to determine.  This  problem is difficult  to  avoid,
as  most  environmental samples contain unknown concentrations of  organic
compounds.   To check if the analyte  is  being  lost somewhere  during the
processing,  a step-by-step examination  of the method  using  a spiked blank
 is  necessary,  with measurements  of the  analyte at each step.

      Emulsions, colored extracts,  or  unusual chemical  behavior  of a  sample
 should be noted  and considered  when reviewing  results.  Modification of
a method by an experienced chemist may  alleviate  some problems (e.g.,  substi-
                                     248

-------
                                                           Bioaccumulation
                                      Priority Pollutant Organic  Compounds

tuting  ethyl  ether for the  usual solvent can sometimes >ielp emulsions  with
tissue samples).  All deviations  from specified methods snould be documented
in logbooks and discussed in  the  data report.

     Results that fall outside  the established calibration curve are suspect
until linearity of  response  can  be  shown at  that concentration  or until
the  extract is  diluted appropriately and reanalyzed.  Extreme values may
necessitate re-extraction of  a  smaller  sample  size.  Extremely high concen-
trations of organic compounds  may saturate the extraction capabilities
of the method and may necessitate re-t .traction of a  smaller sample  size
or use of a more appropriate  method.  Chromatographic  interference by natural
substances present  in the tissue  (e.g.,  fatty  acids) may require additional
cleanup.

     Data  Quality  and  Reporting—A data summary  for each  sample should
be submitted. All  data should  be reported as  ug/kg wet weight to  a minimum
of two significant  figures.  Data should not be blank-corrected.  Spike
and  surrogate recoveries, relative percent difference between duplicates,
and  blank results  (ng/sample)  should  also  be submitted.  The following
additional  information  is needed  for each sample to  allow  independent QA
review:

      0     Sample weight extracted

      •    Final  volume  of extract

      •    Amount of extract  injected

      •    Instrument detection limits

      •    Detection limit for each compound

      •    Blank associated with  the sample

                                     249

-------
                                                             Bioaccumulation
                                        Priority Pollutant Organic Compounds


     •    Deviations from the prescribed method


     t    Problems associated with analysis.


     For a more thorough QA review, additional documentation (e.g., chromato-
grams, computer listings) may be requested.
                                      250

-------
                              Demersal Fishes and Large Macroinvertebrates

Demersal  Fishes  and Large Macro invertebrates

Field Procedures—

     Collection--After the trawl assembly has been brought on board, the
cod-end choker  can be  released  and the catch emptied  into some kind  of
collection vessel  (e.g., tub,  tray,  bucket,  special table).  It  is reconnended
that a collection  vessel  be used to prevent  organisms  from escaping and
to prevent possible contamination of tissue-contaminant specimens by exposure
to the deck surface.

     If organisms from the trawl  catch  are  to be  subsampled  for tissue
contaminant analyses, these individuals should be noted and removed  from
the  sample before the  remainder is processed  for  ecological information.
Care should be taken  to minimize  potential  contamination  of these specimens.
For  example, the deck  and all  collection containers  or equipment should
be washed before and  after each haul.  Smoking should not be  allowed  during
sampling and all crew members  should wear  clean gloves when handling the
organisms.  It  is  also  helpful if the skipper  positions  the vessel  so  that
stack gases move away from  the fantall.   Additional precautions are  given
 in the bioaccumulation  sections of this document  that describe  the  detailed
collection methods to be used for specific  kinds  of  tissue analyses.

      Processing--Once tissue-contaminant specimens  are removed, the remainder
of the catch can  be processed without concern  for contamination.   All  organisms
should be  Identified and counted.   Identifications  should be to  the  species
 level,  if possible.   It is advisable to  have  a collection  of taxonomic
 keys to  local  fauna on board  the  sampling vessel. For fishes, scientific
 and  cannon nanes should conform to  those recommended by Robins  et al.  (1980).
 If the  identity  of a species is in doubt, several   representative specimens
 should be frozen or fixed with 10-percent buffered formalin  and  identified
 by an expert  after the cruise.
                                     251

-------
                              Demersal  Fishes  and  Large Macroinvertebrates

     Following Identification, the size of each  individual should be measured.
For fishes, it is  recommended that total  length  (TL)  be  determined to the
nearest  millimeter (mm).  Total length  is  the length from the anterior-
most part of the  fish to the tip of the longest  caudal  fin rays.   Two kinds
of  total  length can be measured (Anderson  and Gutreuter 1983).  Maximum
TL  is determined when the lobes of  the  caudal   fin  are  compressed dorso-
ventrally, whereas natural  TL is measured  when  the caudal fin is in  its
natural state.  To be consistent with the :onvention  used  by most  fishery
investigations in the U.S.,  maximum TL  should be  measured (Anderson  and
Gutreuter 1983).

     For  fishes  that  occur  in  relatively large numbers,  individuals  may
be classified into  size classes rather than measured  exactly.  This procedure
can  reduce processing time considerably.   If size  classes  are used,  it
is recommended that they differ by no more than 1  cm each.  Furthermore,
it  is  recommended that each size class be measured  from one half-cm value
to the next.  For  example, a 10-cm  size would  include  fish from  9.50  to
10.49 on.

     In  some cases,  erosion  of the caudal  fin in a substantial  segment
of  a population may require that a measurement other  than total  length
be  used for  affected  individuals.  If this occurs,  it  is  recommended that
maximum standard  length  (SL) be  used as a  substitute.  Standard  length
is  the length from the anterior-most  part of the  fish  to the posterior
end of the hypural bone.  Anderson and Gutreuter (1983) state that  in practice,
SL  may be measured to some external feature such  as the last  lateral line
scale, the  end of the fleshy caudal  peduncle,  or the midline  of  a crease
that forms when  the tail is bent  sharply.  Standard length can be related
to  total  length by developing  a regression relationship between  these  two
measures  for  a sample that covers the complete  lengch range observed in
the population.

     As  each individual  is  measured,  it  should also be  scanned  carefully
for grossly visible external abnormalities.  Some common external abnormalities

                                    252

-------
                               Demersal  Fishes and Large Macro invertebrates

found  in fishes captured  in polluted areas are listed in Table 7.  Recent
literature regarding  this subject  is  reviewed  by Sindermann (1979).   At
least one person on board the sampling vessel should  be capable of diagnosing
the abnormalities listed  in Table 7.  If there is  any doubt as to the identity
of an  abnormality,  then  one,  or  preferably several, specimens afflicted
with  the disorder should  be frozen or fixed with  10-percent buffered formalin
and  taken to an expert for accurate diagnosis. Once disorders have been
confirmed, it is helpful  to maintaizn reference  specimens on the  sampling
vessel  or in a shore-based laboratory.

     Perhaps  the most  frequent  error associated with  recognizing external
abnormalities is mistaking net-related damage as  integumental  lesions.
This problem occurs  most frequently when diagnosing  ulcers  and  fin erosion.
Net-related damage can result from  surface abrasion  as captured  fishes
are dragged along the  bottom in  the cod end or as  .fishes  swim  against  the
net in an attempt to escape.  Net-related damage can also  result from captured
fishes  biting each  other.  Whenever a suspected lesion appears  to be  newly
formed  or is actively bleeding, net-related damage should  be suspected.

Laboratory Procedures--

     There are no laboratory procedures for demersal  fishes  and megainverte-
brates, as all processing will occur  in the  field.
                                     253

-------
TABLE 7.   SOME COMMON  EXTERNAL  ABNORMALITIES OBSERVED
            IN FISHES  FROM POLLUTED AREAS
              Integumental  Lesions
              Fin  erosion (fin  rot)
              Lymphocystis
              Ulcers
              Neoplasms  (tumors)

              Skeletal Malformations

              Scoliosis
              Lordosis
              Cranial  compresc,  in (pugheadedness)
              thv.rfism
              Deformed  fin  rays (bent fins)

              Pigmentation  Anomalies

              Ambicol oration

              Parasites

              Copepods
              Isopods
              Nematodes
              Leeches
                           254

-------
                               REFERENCES


Anderson, R.O.,  and  S.J. Gutreuter 1983.   Length, weight,  and associated
structural  indices,  pp. 283-300.   In:   Fisheries  Techniques.  L.A. Nielsen
and O.L. Johnson (eds).  American  Fisheries Society,  Bethesda, MD.

Birkett, L., and A.O.  Mclntyre.   1971.   Treatment and sorting samples.
pp. 156-168.   In:  Methods  for Study of Marine Benthos.   N.A.  Holme and
A.D. Mclntyre (eds).   IBP Handbook No.  16.  Blackwell  Scientific Publications,
Oxford, UK.

Bcrdner, R., J. Winter, and P.  Scarpino.  1978.  Microbiological methods
f r - monitoring the  environment:   water and wastes.   EPA-600/8-78-017.
Environmental Monitoring and  Support Laboratory, U.S.  Environmental Protection
Ag.»ncy, Cincinnati,  OH.  337  pp.

Buchanan, J.B., and  J.M.  Kain.   1971.   Measurement  of the physical and
chemical environment,  pp.  30-52.   In:   Methods  for the Study  of  Marine
Benthos.  N.A. Holme and A.O. Mclntyre (eds).  IBP Handbook No. 16.  Blackwell
Scientific Publications, Oxford, UK.

Oisalvo, L.,  H.E.  Guard, and  N.D.  Hirsch.  1977.  Assessment  and  significance
of  sediment associated oil  and grease in  aquatic  environments.   Technical
Report  0-77-26.   Naval  Biosciences Laboratory, Naval  Supply  Center, Oakland,
CA.  145 pp.

Elliott, J.M.  1971.   Some  methods for  the statistical analysis of  samples
of benthic invertebrates.  Scientific  Publication No. 25.  Freshwater Biological
Association, Ferry House, UK.   148 pp.

Fauchald, K.   1977.  The polychaete  worms; definitions and keys  to the
orders, families,  and genera.   Science Series 28.   Natural History Museum
of  Los  Angeles County.  Los Angeles, CA.   188 pp.

Folk,  R.L.   1968.   Petrology of sedimentary rocks.  University of Texas,
Austin, TX.  172 pp.

Glbbs,  P.J.. and  J. Mathews.  1982.   Analysis  of experimental  trawling
using  a miniature otter trawl to sample  demersal fish in shallow estuarine
waters.  Fish.  Res. 1:235-249.

Gonor,  J.J.,  and  P.P. Kemp.  1978.  Procedures for  quantitative ecological
assessments in intertidal environments.  EPA-600/3-78-078.  U.S. Environmental
 Protection Agency, Corvallis, OR.   104 pp.

Gosner, K.L.  1971.  Guide  to identification of marine  and estuarine inver-
 tebrates.  Wiley-Interscience, New York, NY. 693  pp.

 Green, E.,  0. Schnitker.   1974.   The  direct  titration of water-soluble
 sulfide in estuarine muds of Montsweag  Bay, Maine.  Mar. Chem. 2:111-124.

                                     255

-------
Greenberg, A.E., R.  Trusseil, and L. S.  Clersceri  (eds).  1985.   Standard
methods for examination of waste and  wastewater.  American  Public Health
Association, Washington, DC.  1268 pp.

Hiatt, M.H.  1981.   Analysis of fish  and  sediment  for  volatile priority
pollutants.  Anal.  Chem. 53:1541-1543.

Hinricks, D.J.  1979.   Inspector's  guide:  to be used  in  the evaluation
of municipal wastewater treatment plants.  Prepared for Municipal Operations
Branch, Office of Water Program Operations, U.S. Environmental Protection
Agency, Washington,  DC.

Holme, N.A.  1971.   Macroinfauna sampling,  pp. 80-130.  In:  Methods for
f.-cudy  of Marine Benthos.  N.A. Holme and A.D. Mclntyre  (eds).   IBP Handbook
No.  16.  Blackwell  Scientific Publications, Oxford, UK.

Howmiller,  R.P.  197 .  Effects of preservatives on weights  of  some common
macrc'^nthic  invertel-rates.  Trans Am. Fish Soc. 4:743-746.

Kitchens, J.F., R.E. Casner, S.S. Edwards,  W.E. Harward III, and B.J. Macri.
1976.   Investigation of  selected  potential environmental  contaminants:
formaldehyde.  EPA-560/2-76-009.   U.S.  Environmental  Protection  Agency,
Washington, DC.

Lappalainen, A.,  and P.  Kangas.   1975.   Littoral  benthos of the northern
Baltic Sea.   II. Interrelationships  of wet, dry, and  ash-free weights of
macroinfauna  In the Tvarminne area.   Int. Rev. Gesamten. Hydrobiol. 60:297-312.

Mearns,  A.J., and J.M. Allen.   1978.   Use of small  otter  trawls  in  coastal
biological  surveys.  EPA-600/3-78-083.   U.S. Environmental Protection Agency,
Corvallis,  OR.  33 pp.

Metcalf 4  Eddy, Inc.  1979.  Wastewater  engineering:  treatment/disposal/reuse,
Second Edition.  McGraw H111  Book Company, New York,  NY.   920  pp.

Mills, E.L.,  K.  Pittman, and B.  Munroe.  1982.   Effect  of preservation
on the weight of marine benthic  invertebrates.   Can. J.  Fish.  Aquat.  Sci.
39:221-224.   -

Plumb,  R.H., Jr.   1981.   Procedures for handling and chemical  analysis
of sediment and water samples.  Technical  Report EPA/CE-81-1.   U.S.  Environ-
mental  Protection Agency/U.S.  Corps of Engineers Technical Committee on
Criteria for Dredged and  Fill  Material, U.S.  Army Waterways Experiment
 Station, Vicksburg, MS.  471 pp.

 Robins,  C.R.,  R.M. Bailey, C.E.  Bond, J.R.  Brooker, E.A. Lachner,  R.N.  Lea,
 and W.B. Scott.   1980.  A  list of common  and scientific names of fishes
 from  the  United States and Canada.  Special  Publication No. 12.   American
 Fisheries Society,  Bethesda, MO.   174 pp.

 Schwinghamer, P.  1981.   Characteristic  size distributions of integral
 benthic connunities.  Can. J. Fish. Aquat. Sci.  38:1255-1263.


                                      256

-------
Shepard, P.P.   1954.  Nomenclature  based  on  sand-siU-clay ratios.   J.  Sedi-
ment. Petrol.  24:151-158.

Shepard,  P.P.  1963.  Submarine  geology.  (Second  edition).   Harper and
Row, Publishers.   New York, NY.   557  pp.

Sindermann, C.J.  1979.  Pollution-associated diseases and  abnormalities
of fish and shellfish:  a review.   Fishery  Bull. 76:717-749.

Smith,  R.I.,  and J.T. Carlton  (eds).  1975.  Light1 s manual:   intertidal
invertebrates  of  the central  California  coast.   University of  California
Press, Berkeley,  CA.  716 pp.

Stfiphenson, M.O., M. Martin,  S.E.  Lange,  A.R.  Flegal, and J.H.  Martin.
19/9.  California mussel watch 1977-1978.   Volume  II:  Trace metals concen-
trations  in the  California  mussel, Myti1 us  californianus.   SWRCB  Water
Quality Monitoring Rer >rt No. 79-22.   Sacramento, CA.  110 pp.

Stephen-son, M.O.,  S.L. Coals, M.  Martin,  and J.H. Martin.  1980.   California
state mussel watch 1979-1980.  Trace  metal  concentrations in the  California
mussel , Mytil us  californianus.  and the  bay mussel, Mytilys  edulis,  along
the California coast and selected harbors  and bays.  SWRCB Water Quality
Monitoring Report 80-8.  Sacramento,  CA.  95 pp.

Stofan, P.E.,  and G.C. Grant.  1978.   Phytoplankton  sampling in quantitative
baseline  and  monitoring  programs.   EPA-600/3-78-025.  U.S.  Environmental
Protection Agency, Corvallis, OR.   27 pp.

Strickland, J.O.H., and T.R.  Parsons.  1972.   A practical handbook of seawater
analysis.   Bulletin No. 167.   Fisheries  Research Board  of  Canada, Ottawa,
Canada.  310 pp.

Striplin, P.L.,  and S.H. Maupin.  1982.   Custom-designed sieving stations
for small research vessels.  Task report to  the U.S.  Environmental Protection
Agency.  Corvallis, OR.

Swartz, R.C.  1978.  Techniques for sampling and analyzing the marine macro-
benthos.  EPA-60073-78-030.  U.S.  Environmental Protection Agency, Corvallis,
OR. 27 pp.

Tetra Tech.  1982.  Design  of  301(h) monitoring programs  for municipal
wastewater discharges to marine waters.   EPA 430/9-82-010.  Office of  Water
Program Operations, U.S. Environmental Protection Agency, Washington,  OC.
135 pp.

Tetra Tech.  1985a.  Bioaccumulation monitoring  guidance:  2.  selection
of target  species and  review of  available bioaccumulation data.   Final
program document prepared  for the  Marine Operations  Division, Office of
Marine  and Estuarine  Protection, U.S.  Environmental  Protection  Agency.
EPA  Contract No. 68-01-6938.   Tetra Tech, Inc., Bellevue, WA.  52 pp. +
appendices.
                                     257

-------
Tetra  Tech.  1985b.   Bioaccumulation monitoring guidance:   3.   recommended
analytical detection  limits.  Final program document prepared for the Marine
Ooerations Division,  Office of Marine and Estuarine Protection,  U.S. Environ-
mental Protection Agency, Washington,  DC.  EPA Contract No.  68-01-6938.
Tetra Tech, Inc., Bellevue, WA.  23 pp.

Tetra Tech.  1985c.   Summary of U.S. EPA-approved methods, standard methods,
and  other guidance  for 301(h) monitoring variables.  Final  program document
prepared  for the Marine Operations Division, Office of Marine and Estuarine
Protection, U.S. Environmental Protection Agency.  EPA Contract No. 68-01-6938.
Tetra Tech, Inc., Bellevue, WA.  18 pp.

Tetra  Tech.  1986a.   Analytical  methods for  EPA  priority pollutants  and
301(h) pesticides in  estuarine and marine sediments.  Final  program document
prepared for the Marine  Operations Division, Office of Me.rine and Estuarine
Protection, U.S.  Environmental Protection Agency.  EPA Contract No.  68-01-6938.
T?tra Tech, Inc., Bellevue, WA.

1»tra  Tech.  1986b.   Bioaccumulation monitoring guidance:   '-.   analytical
methods  for U.S. EPA  priority pollutants and 301 (h) pesticides  in  tissues
from estuarine  and  marine organisms.   Final  report prepared  for Marine
Operations Division, Office of Marine and Estuarine Protection, U.S. Environ-
mental  Protection Agency.  EPA Contract No. 68-01-6938.   Tetra Tech,  Inc.,
Bellevue, UA.

Tetra  Tech.  1986c.   Evaluation of coastal survey positioning methods  for
nearshore marine and  estuarine waters.   Final  report prepared  for  Marine
Operations Division,  Office of Marine and Estuarine Protection,  U.S. Environ-
mental Protection Agency.  EPA Contract No. 68-01-6938.   Tetra  Tech,   Inc.,
Bellevue, WA.  54 pp. + appendices.

Tetra  Tech and  American Management Systems.   1985.  ODES data submissions
manual.   Prepared for U.S. EPA by  Tetra Tech,  Bellevue, WA and American
Management Systems,  Washington, DC.

U.S. Department of  Transportation.  1984.  Hazardous materials  regulations.
U.S. DOT, Washington,  DC.  Federal Register, Vol. 49, Chapter 1, Subchapter C,
pp.  52-792.

U.S. Environmental  Protection Agency.   1977.   NPDES compliance  sampling
manual.   Enforcement  Division, Office of Water Enforcement Compliance  Branch,
Washington, DC.  139 pp.

U.S. Environmental  Protection Agency.  1978  (revised 1983).  NEIC  policies
and  procedures.  EPA-330/9-78-001-R.  National  Enforcement  Investigations
Center,  Denver, CO.

U.S. Environmental  Protection Agency.  1979a.  Handbook  for analytical
quality control  in water and wastewater  laboratories.  National Environmental
Research Center, Cincinnati, OH.

U.S. Environmental Protection  Agency.   1979b  (revised March, 1983).   Methods
for  chemical analysis of  water and  wastes.  EPA  600/4-79-020.  Environmental
Monitoring  and  Support Laboratory,  Cincinnati,  OH.

                                     258

-------
U.S. Environmental Protection Agency.   1980.   Interim  guidelines and specifi-
cations for preparing quality assurance project  plans.  Office of Monitoring
Systems and Quality Assurance, Washington DC.   18  pp.

U.S. Envronmental Protection Agency.   1982.  Handbook  for sampling  and
sample preservation of water and wastewater.   EPA-60074-82-029.  Environmental
Monitoring and Support Laboratory, Cincinnati, OH.   402 pp.

U.S.  Environmental  Protection Agency.   1983a.   Environmental measurement
method performance data  for  establishing achievable data  quality goals.
Environmental  Monitoring Systems Laboratory,  Research  Triangle Park, NC.

U.S. Environmental Protection Agency.   1983b.   Guidance for preparation
of  combined work/quaHty assurance  project plans  for  water monitoring.
Office of Water Regulations and Standards, Washington, DC.  33 pp.

U.S.  Environmental  Protection Agency.  . '84.  Guidelines establishing test
procedures for  the analysis cf pollutants.  U.S.  EPA, Washington, DC. Federal
Register, Vol. 49, No. 209, pp. 43234-43436.

U.S. Environmental Protection Agency.   1985.   U.S. EPA Contract Laboratory
Program  - statement of work for inorganic analyses, multi-media multi-
concentration.  U.S. EPA, Washington,  DC.

U.S.  Navy.  1968.   Instruction manual  for obtaining oceanographic data.
Publication No. 607.  U.S. Naval Oceanographic Office, Washington, DC.

Wiederholm, T., and L. Eriksson.  1977.   Effects of alcohol preservation
on the weights of some benthic invertebrates.  Zoon 5:29-31.

Word,  J.Q.  13 February  1985.   Personal Communication  (discussion with
Dr. Gordon R.  Bilyard).  Evans-Hamilton, Inc., Seattle, WA.
                                     259

-------
                                GLOSSARY


a*b - Notation for  acid and alkali  solutions.   In additive volumes  (a+b),
tne  first number, a,.refers  to the volume of a concentrated  reagent; the
second number, b,  refers  to the volume  of distilled water required for
Dilution.

Accuracy - The closeness of a measured or  computed value to its  true value.

Aliquot - A divisor that divides a  sample  into a number of equal  parts,
leaving no remainder; a sample resulting from such a divisor.

Ambient - Surrounding, encircling.

Analyte - The specific component measured  in a  chemical analysis.

APHA - American Public Health Association.

Apparent Particle Size Distribution  -  Distribution comprised of both the
inorganic and  organic  particles in  a  sample (i.e., organic  material is
not oxidized).

Archive • A repository of evidence or information.

Asymptote - A  line considered  a limit to  a curve  in the  sense that the
perpendicular distance  from  a moving point on the  curve to the line approaches
zero as the point moves an infinite distance from the origin.

Auto-Injection  -  An automated introduction of a sample into an instrument
for  analysis.  Usually  used  for  extracts analyzed  by gas chromatography.

Background Correction  Factor  - A  number  that adjusts the analyte  signal
for  interfering matrix  effects  in flame and graphite furnace  AA analysis
(usually done automatically by the  instrument itself through the  use of
continuous deuterium  lamp).

Bacteriostasis • The  arrestment  or  inhibition of bacterial growth and  repro-
duction.

Batch - Usually refers  to the number of samples  that can be prepared  or
analyzed  at one  time.   A typical batch  size is 20 for extraction of organic
com pounds.

Benthic  - That  portion of the marine environment  inhabited by organisms
that live permanently in or  on  the  bottom.

Bioaccumulation - The accumulation of chemical substances in the tissues
of organisms.
                                      260

-------
Biomass  -  The weight of  living material  in all or  part of an organism,
population, or community.   Commonly  expressed  as  weight per  unit area,
a biomass density.

Blank-Corrected  • The concentration of an  analyte adjusted for the concentration
of that analyte  in the blank.

Bow Wake - The pressure wave that forms below a solid object as  it is lowered
through the water column.

Bridle - The line that links each otter board with the  hydrowire.

Calibration - The systematic standardization  of the graduations of the
quantitative measuring instrument.

Carryover - Contamination  arising from previous  analysis or extraction
of a standard or highly contaminated sample.

Chain-of-Custody - The procedures and forms used to  trace and  record environ-
mental samples through all stages of collection,  shipping,  analysis, and
final disposition.

Chelation - The  addition of organic complex ing agents,  such as EDTA (ethylene-
diaminetetraacetic  acid)  that preferentially  bind with metals,  thereby
reducing the exposure and possible toxicity to organisms (e.g.,  bacteria).

Chief Scientist - The person in charge of the  sampling team  on  a  research
vessel.

Cleanup - The removal of co-extracted compounds that may cause interference
from a sample extract.

Cod  End - The mesh bag  at  the back  of  an  otter trawl  in which the  catch
is collected.  It is often of smaller mesh size  than the  remainder of the
net.

Coefficient of Variation - The standard deviation expressed as a percentage
of the mean.

Co-Extractive - Materials other than analytes of interest that are extracted
along with the analytes.  These can be sources of interference.

Composite Sample - A sample composed of two or more grab samples or increments.

Contingency Plans -  Procedures to be followed when those that were  planned
originally cannot be carried out.

Copenhagen Seawater  - Seawater of known chlorinity, an international  standard
for  salinity determinations.

Corrective Action - Measures taken to remove, adjust, remedy, or counteract
a malfunction or error so that a standard or required condition is met.
                                     261

-------
Ooss-Contamination  - Contamination  of  a  sample or  sample extract from
ejoosure  to another sample or sample extract, usually  of higher concentration.

Demersal - Living on or near the bottom of the sea.

Desiccation - T^e thorough drying of a sample by removal of moisture.
Detection Limit  - The smallest concentration of some component of interest
tnat can be measured by a single measurement with a stated  level of confidence.

Digestion - In preparing samples for analysis of metals, an acidic solution
added to break organometall ic bonds,  freeing  the metals  for analysis  by
atomic absorption or  atomic emission spectrophotometry.

Distillation - The vuporization of a liquid mixture with subsequent collection
o*  components by differential cooling to condensation.

Duplicate Analysis - A second analyi  s made  on the  same (or  identical)
sample of material to a_3ist in. the evaluation of measurement variance.

Effluent  -  Something that flows out.   For example,  the  liquid material
discharged by sewage  treatment plants.

Electrolyte - A substance that dissociates into  ions  In  solution or  when
fused, thereby becoming electrically conducting.

Emulsion  - A suspension of  small globules of one liquid  in  a second  liquid
with which the first  will not mix.

Epibenthic - Residing primarily on the sediment surface.

Equilibrium  -  The state  of a reaction in which its  forward and reverse
reactions occur at equal rates so that the concentration  of the reactants
does not  change with time.

Extraction  - A method of separation in which a solid  or  solution  is contacted
with a  liquid  solvent to transfer one or more component  into the  solvent.

False Positive - A  positive measurement of  an analyte* not attributable
to  the  sample.

Field Blank  - An empty  container or uncontaminated  representative matrix
carried  through the field  routine in the same manner  as  a sample.

Filter  Blank - An unused filter extracted and analyzed  in  the  same manner
as  filters  used  to collect  samples.  The filter should be prepared  in the
 same manner as sample filters.

 Fixation - The process  of putting  something  into a stable or  unalterable
 form.

 Footrope - The line that forms the front  edge of the bottom  of an  otter
 trawl.   Weight is often added  to  it to keep it on  the bottom  when being
 towed .

                                     262

-------
Formalin - A trademark for a 37 percent by weight aqueous  solution of formal-
dehyde with some methanol.

Formazin - A polymer used as a reference standard suspension  for turbidity
measurements.

Grab  Sample - 1) see  increment, or 2)  a sample of bottom  sediment collected
by a grab sampler.

Gravimetric - Of or pertaining to measurement by weight.

Headrope - The line that  forms the front edge of the top of  an  otter trawl .
Floats are attached to it to hold the net open when being  towed.

Keadspace - The airspace  between a collected sample and  the  container  lid.

Hydr. wire - The cable  used  to deploy equipment over the side of a vessel.
It usually is attached to a winch at one end and to the  piece  of  equipment
at the other end.

Hygroscopic - Readily  absorbing moisture, as from the atmosphere.

Increment - An individual  portion of material collected by  a  single operation
of  a sampling  device,  from parts  of a lot  separated  in time  or space.
Increments may be either  tested individually or  combined'(composited) and
tested as a unit.

In Situ - In something's  original place.

Interelement Correction  Factor - A number that  adjusts the analyte signal
for  the  interfering effects of other elements present in  a sample  undergoing
ICP  analysis (usually  done automatically by the  instrument itself).

Interference -  A  substance present in the sample that impedes the  accurate
measurement of an analyte of interest.

Interference Check Sample -  A solution contaminating both interfering and
analyte elements of known concentration that can  be used to verify  background
and  interelement correction factors  in ICP analysis.

In Vivo  - Within the living organism.

Isobath  - A contour of constant depth.

Macroinvertebrate - An invertebrate  retained by  a sieve having a  mesh  size
of 1 mm.

Matrix Spike Compound  - A known  amount of  an analyte added to a sample,
usually  prior to extraction or digestion.

Method Blank -  The contamination  by the  analyte from all sources  external
to the sample.  The blank value  is determined by proceeding through all
phases of extraction and analysis with no addition of sample.

                                     263

-------
NBS - National Bureau of Standards.

Offshore - In this document,  refers  to all aspects of the receiving environment
for effluents (i.e., water,  sediirent, and organisms).

Otter  Board - A  flat board that  is attached to each side of the front end
of an otter trawl. Its planing  action when  being towed  holds the mouth
of the net open.

Oxidation - An increase in positive  valence or a decrease in negative valence
by the loss of electrons.

Penetration Depth - The maximum  depth below  the sediment surface  that -
grab sampler achieves during  a  single cast.

Performance Samp". • - A sample  or  solution  with  known concentrations of
anaiytes of interest, submitted  to a  laboratory for the purpose of evaluating
the performance of that laboratory.

Phi Value - A measure of particle size commonly used by  geologists.  A
phi value is equal to the negative  logarithm  (base 2) of  the diameter of
a particle expressed in millimeters.

Population - A generic term denoting  any finite or infinite collection
of individual  things, objects, or  events; in the broadest concept, an aggregate
determined by some property that  distinguishes things that do and  do not
belong.

Precision - The degree of mutual  agreement characteristic of independent
measurements as the result of  repeated application of the  process under
specified conditions.

Preservation - Maintenance in an  unaltered form.

Preventive Maintenance - Procedures conducted routinely  to ensure that
equipment continues to operate  properly.

Primary Standard -A substance  or artifact, the value of which can be accepted
(within  specific Units) without question when used to establish the value
of the same or related property of another material.

Priority Pollutant - Those toxic  pollutants  defined by  the U.S.  EPA  in
1976 that are the primary subject  of regulation of the  Clean Water Act.
A  list of these  substances  can  be  found in the Code of Federal Regulations
Vol. 40, Section  401.15.

Purge  -  The removal  of volatile  organic compounds from the sample matrix
for analysis.

Quality - An estimation of acceptability or suitability for a given purpose
of an object, item, tangible, or  intangible thing.
                                    264

-------
Quality  Assurance - A system of  activities to provide to  the  producer or
user of a product or a service the  assurance that it meets defined  standards
of quality.
                                 V
Quality  Assessment • The overall  system of activities wh-ose purpose is
to provide assurance that the quality control  activities are  being done
effectively.   It  involves a continuing evaluation  of performance  of the
production system and the quality of the products produced.

Quality  Control  - The overall system of activities whose purpose  is to
control the quality of a product  or  service so  that it meets the  needs
of users.   The aim is to provide  quality that is satisfactory,  adequate,
dependable,  and economic.

Quantification -  The determination  or expression of the number  or  amount
of something.

Reagent  -  A solvent or other chemical used during  sample  p  »paration or
analysis.

Recovery -  The amount of an analyte detected relative to the amount  added
(e.g., spike) or known to be  present (e.g., standard reference  material).
Usually expressed as a percentage.

Reduction  - A  decrease in positive  valence by  the  loss of electrons or
an increase  in  negative valence by  the gain of electrons.

Reference  Area -  A station or group of stations  with which  potentially
impacted stations are compared  to determine the degree  of impact.   Ideally,
the reference area represents unaltered  background conditions.

Reference Collection - A group of preserved organisms of known and  verified
taxonomic  identity that is used  as the standard for  comparison for  future
taxonomic identifications.

Relative Percent  Difference - Difference of  two  measurements xj and X2,
divided by the  mean of the measurements, multiplied by  100.

Relaxation - Reduction of muscular  or nervous tension.

Replicate - A counterpart of  another,  usually referring to an  analytical
sample or  a measurement.  It is  the  general  case for which duplicate is
the special  case consisting of two  samples or measurements.

Reproducibility - The ability to produce the same results for  a measurement.
Often measured  by calculation of relative percent difference  or coefficient
of variation.

Resection -  The surgical removal of part of an organ or structure.

Sample - A portion of a population  or  lot.  It may consist of  an  individual
or groups of individuals.  It may refer  to objects, materials, or to measure-
ments,  conceivable as part of  a  larger group that could  have been  considered.


                                    265

-------
Sample Integrity  - The unaltered  composition of a sample.

Sample Matrix • The material  in which the analytes of  interest are found
(e.g., water, sediment, tissue).

Sample  Tracking  - Monitoring the  course 0" samp1-?* through all  phases of
laboratory analysis.

Scope - The length of hydrowire used  when  towing an otter trawl.

Secondary Standard - A standard whose value  is  based upon  comparison with
some primary standard.

Seed  -  A population of microbiological organisms added  to a sample  for
BOD analysis because the sample  does not contain a m.irobial  population
sufficient for the needs of the analysis.

Sensitivity • Capability of methodology  or instrumentation  to discriminat
between  samples having differing  concentrations of an analyt* .

Significant Figure -A figure(s)  that remains to a number or decimal after
the ciphers to the right or left  are  cancelled.

Sort  - To separate benthic organisms  from  the inorganic  and plant material
that  are collected in sieved grab samples.

Spike •  The addition of a known amount of  an  analyte or internal standard
to a  sample.

Split - A replicate portion or sub-sample of a total sample obtained  in
such  a manner that it is not believed to  differ significantly  from other
portions of  the sane sample.

Standard -  A substance or  material,  the  properties of which are believed
to be known  with sufficient accuracy  to  permit its use  to evaluate  the
same property of another.   In  chemical measurements, it  often describes
a  solution of substance,  commonly prepared by the analyst, to establish
a  calibration curve or the analytical response  function  of  an instrument.

Standard Reference Material - A material or  substance one or  more properties
of which are sufficiently  well established  to  be used for the assessment
of a method  or the calibration of an apparatus.

Sterilization - Reaoval  of all bacteria  and  other microorganisms from an
object.

Subsample  - A portion  taken  from  a sample.   A laboratory sample may be
a subsample of a field sample.  Similarly, a test  portion may be a subsample
of a laboratory  sample.

 Surrogate Spike  Compound - A known amount of a  compound, with characteristics
 similar to that  of an analyte, added to a sample prior  to  extraction.  This
 compound  can be used  to  estimate  recovery of analytes of interest.  Also
 called  "recovery internal  standard."

                                    266

-------
Tare  -  The  waight of a container  or  wrapper  that  is deducted from the gross
weight to obtain  net weight.

Taxon  -  A  group of organisms  constituting  one of the categories or formal
units in taxonomic clarification, such  as  phylum, class,  order,  family,
genus, or species.

Taxonomy -  The theory, principles, and process of classifying organisms
in established  categories.

TFE  -  A  plastic tetrafluoroethylene polymer composed of very long chains
of CFg units.   Commonly known by the  trade name Teflon.

Titration -  The process or  method of determining the  concentration  of a
substance in solution by adding  to it a  standard reagent of known concentration
in carefully  measured amounts until a reaction  of definite and known proportion
is completed,  as shown by  a color change  c   by electrical  measurement,
and then calculating the unknown concentration

Trace - Very small quantity of analyte in the sample.

True Particle Size Distribution  - Distribution comprised  only of  inorganic
particles after organic material is oxidized  completely.

Volatile Organic  Compounds - Organic  compounds with high  vapor pressures.
In this document it refers to the 29  U.S. EPA priority pollutants considered
as volatiles (e.g., benzene).

Winnow - To separate different constituents  of a substance by means of
a current of air or water.

ZID - Zone of initial dilution for an effluent discharged  into the environment.
                                     267

-------
APPENDIX A

-------
                 APPENDIX A
RECOMMENDFD METHODS FOR METALS IN EFFLUENT

-------
                    TABLE  A-l.   LIST OF TEST  PROCEDURES  APPROVED  BY U.S.  EPA
                                 FOR INORGANIC  COMPOUNDS  IN  EFFLUENT
        Note:   This table  is an exact  reproduction  of  Table IB (U.S.  EPA  1984),
                                                                 •«• IIIUHOB No. « MOM
                                    EM H7»
                                                                     ASTM
                                                                                    USCS>
                                                                                                     Ora
             mg/L.
       aT-lT-i -p" •———-
       » pH 41 mm*
sw.i-
        Tow • mot. Ogiinnr •
                                 L1-
                               JBM..
•02144)

403

SOX
                                                               01M7-
                                01087(8).
                                                i-1030-rt
        W> NX ng/L i  MI MMW W
                                               4I7A_
                                               4I7«
                                               4170
                                                               014»-7M)).
                                                                               M9I3-7I
        TOM*, man.
                                               303A.
                                               304..
                                                                                I-30U-7I
                                               3071.
                                                                01972-7IU
        c«' «g/t.
        •L pf j
        •TcM'. MQO.
                3O4.
                                •Ml.
                                W*.
                                                                                  H-7«.
                                                                P iU*
                                                                                                    >oor«
                                                                                                «*MWd>007<
                                                                     2007*
t e.
10
11 Brand*. "0/«- Tc
 13.
        -Taw*
   OEDTAI
                                213J
                                H3J
                                J14I-
                                11U
                                                S07.
                                                40M.
                 3101.
                 300*.
                                                31 tC
                                                                OSU7.7MO.
                                                                 0111
                                                                 OS11-77(*.
                                                                                I-3I12-7I
                                                                                                MMied2M7<
                                                                 P 37«
                                                                                                     200.7'
                                                        A-l

-------
TABLE  A-l.   (Continued)
                          I-
                                EM it?t

                                                                      *3Tli

                                                                0912-Wm
                                                               i M12-47IA)
                                                                DS12-«7 M0*4nl

                   . .  . P 17*tM

                       I'M
 ("I




 P 5541

1

I
                                                                                                     n
                       P M7«


                       M(Md2007<

                                                                                   UIH0.7I
                                                                                                     i,..,
                                                                                                      P a*
                                                                                                       IM7iod2007<
                                                       A-2

-------
        TABLE A-l.   (Continued)
                                       CM l
                                                            u.
                                                                                              USGS-
          01 IpMXpMiM*
40.
                                                                      Oim-7NA>or 01 «»• '  -t
                                                                        nm.            I
   O A* kmm.
3* kot-T mgfl.
                                                     304.
                                                                      OIOM-77 ,
                                                                      (C«0»_
                  iMfMMt.
31
               -row IH NI *§rt-
                                                    31M-
                                   »X-
                                   *Ju
                                   »J_
                                   au
42Merl.
4tro	
41«	
4I7C	
oioak-rriM .. ..


cnStn'."_ ~
       -TOM*, mgn.
                                     L1_
                                   au_
                                                     30Mar309S.
                                                                      Dim T>tAa»B-
                                                                       cmai TJCQ..
           -Toot*.
   » mm tni*i	
M. MOTtaraw-TeW1
                                                    311
                                                    111
                                                                       OH*-77(A)-
                                                                      oiM*.rr tc • Q.
                                                    »•_
       r (•§ HI m»rc
                                                                          •T1..
           I(MN|.«*U
                                                    41K-

                                                    41».
                                                                      OM(7-mA|_
«. «• M
   OAAI
                   •MtlWMk
                                                                                                         I'"I
                                                                                        I-3U1-7I
                                                                                                            Uf •
                                                                                        t-ISU-71
                                                                                        UM47.T1_
                                                                                                  )3>wd
                                                                                                          P SST*
                                                                                                          >n»ea aco r«
                                                      P MM


                                                      HiXOd 200 7'

                                                      P IM*
                                                      It
                                                                                                          It
                                                                                                          P Ml.1
                                                                                                          P 53ft •
                                                                                                          P saa»
                                                          A-3

-------
          TABLE  A-l.    (Continued)
                                                                                     INB.V
        Pn^MW imn ind wviMtf
                                           EPA tf7»
                                                                      ism
                                                                                  ASTM
                                                                                                                        OM>
            -TDW. fin/L.
SI "moum-Tooi •  man. Ogmon
   Or«ww
SI "i»aui  Mia mfl/l_ Grtwnmc.
M n.iax  »*«nrii« mg^L GOMIMM. IWC
U glaum  mniiiiiriBH  ftSS)  mg/L  G>n»
M l»i»au«  MBIHB .  ng/L  vokmnc (MM0
}7 Pin hi  iuam nn . GrtwmMx. UOX ..-
               "*V-
        «a SO.X mg/L
« SUMi M> &  tW(/L
71


Tl
r«ML» mg/j. Bl	
           ntg/L
          WTU NwMemmc	
            Tow.'
   AA draa m»«M»i 	
   « MTWIMIC KMe «o«
71 Cne-Tout-  mg/L
   AA ««a uornere	
             El2kOM| .
                                      MS1«MU
                                      JH.1	
                                      3M.«	
                                                 *<«C S47>
                                                                                                          p»> Mi-a>>
                                                                                                          t'«V
                                                                                                                 MWM30a7<
                                 tl_
                              MU-
                                                 XM.
                                      MMO-70.
                                                        214*.
                                                                           O1MO-71.
                                                                                           »-7a
                                                                    093n-79_
                                                                                                                 M*nu02007*
                                                        MM or]
                                                        304	
                                                                    oiwi-ntoi-
                                                                    ot«n-rr(o_
                                                                                                   7».
                                                         »> M7»
                                                         p p«
                                                                                                                         Ii7«
                                                                A-4

-------
        TABLE  A-l.    (Continued)
                                                                     To» _______
              19 *•>*»•» 01 "weon* SuMoncoo « v»«e ^d too* Soa¥nan>v- US OoBd>tmoiii n no mm US Cmou* SIM Qaon-'
            gi ino>ganc SuMiancn « *•<• ond F^nvi 5ii*m»m.- NW Skeugno. at *» US Gaooocoi Sirwy  T«nnaun o> v»our Aciowm
                                                                               i»>*d|i9B01
        •*• wiormmato* 31 101*1 «•«*« «• uneM » not wor*o Mtert grec tiara A ugoiuon orecoowo « reaped to lotaon* luoowaoa nwanoi *na 10 doiva* onto* arganc « n n i moon ottano. o* •*• IB oaiommamn* tar e*ru« aumom aucfi a* arwnc.  M non* itni«ia  marewy. tmjn. ana nonun I*OM* a "aonrni
**•*•*• •"• '" •• e§»" •* """^ """• «•""* M «""«**'t' «0«a*t muvenono and/or ejutena.	
    HfgBi n my tfg***on prooodw* tor dvla0 aapvoton or yaffln kfnaoo awnc ofioorpoon •noanvB i^dbdod 01 on* ol M ooior aoproMd rotorancoa • otftarom man 01* *DOV* *n* E0*
                                                •« pat weugn • 0 «J moan »"»i«im« *av ^aaoMig N»iMn at in* Mno* «• r«M>«nc*a proeMu>* >v tout m«jit
                                                   or ngnnnn at u« orgr« iwnpK moan "or aw moun mor ao omnoo >or AA iar«i UOMBOX v gjiom* iurnfc.ii
                                                               Seocvomtinc M«noa tar Tree* EMrmm AnvyM o« w«i«r ana vnoun » an*« «i Aeovai C 31 im P«I
    • 1ignj*t omuunon a not raggHl^ UiiioanDtn flau on iicnHnun* «mu«m aamoM* ar* on a»  N*. • :•.

'                                      (C80CX) mm not Do OOMUMO •"» m* noononoi BOO> law • It rnoaonaa tow BOO  Tho oooMn ol in* nmtcaDon MWHQT .a  mi
           ra rocwii Oni ««» • aomoiBjr > panM iprarctfy turn CBOOk « '•»*•• em DM p*mnoo i*eon au» oourao uovio, w*
          an NMml StMofdOK Mwoa^ PmwngEHkim. At* L I97S AmoM kom ANSI. ««» aVotaoof NOB van NT IOC
                                                                Bo. 2990 Coxoao StakalWl 77»40
                      iw p«fl*r tnduon » *• and Sworn mo
                                                                           10019
                                                           .    .                  .
                               »«*cidtS01 MKf. H»I*»OO» at Wtlor Anory**. 1979. MOON OwncM Comotn* . »0 Boi MS. U»«Ml»x* "»»mcidi n EPA MamoBi 33SOJ iCyanaal or «» 2 (cnonaiM or* amoXiod 6» eomKimg n* ro-umoH in* arocny 10 n*
                            IMB IWOT M MOMd OS B» Mm « 2 moKd Do 'ootiead •*< BH DufUr 7 1 tound a UOBIOO 33» 2
                           tid&eMdt MMtod. mauow umod Nurwar J7»-7SWA, Omoor 197*. Ttemcan AMO-AMTfiar H TKMMOH mauuiw Sniora
                                       1009. 1990. MOO» CKomeal Company. PO te 399 LOMUne. CoMroao 90537
                                             H9a Hoen MandDoo* ol woaiovaior AnaMav t—      	
                                             e*ny. PO  Bm 199  Lovoiana Couraao 90S37
Mama. Modiud 9030 Hocn MandDoo* ol woaiovaior AnaMav 1979 oogoa 2-n3 and 2-n7  Mien Cnomcai Comam*  Lo.aiane Cooaoo 9M17
                  --  -   —     ..  -  - .   ao (0537
                                                                                  Cnrornnogriony  jownol ol CnromiiogriBny voi 47 NO  1  00 42<- on aovooua Ouflor ol kooMn moauitaia ana uoum "rffoioa 10 a a»
 « i J nw*»oV* tar *<•• ol MM ODOM 1 nw/L »ink ol UR«M tfouM DO OM*d 10 100 im. or Movtg 40 m. aocn ol 2 M Na.SA and 2M NoOn Sianaaroa ineud DO araovoo « M
 MM mavvvr  Por^vooi ot a*»*r oocw i >nyL W iaconvnanoa0 "* I""*T *> infantry
 i""*——*   HH F—,  jr om limntfi r  •Woar TomuBiMo innygnail Ticion. rigid Momtonmn ana Oiu P>»o«miBDn.  U S  Gadoacal Sun»» T*ennojya 01 *aioi «•
             '   i  i Ooptor 01  1971
           zJw kMOaoMMM 9009 HO* Hondboo* ol W*» Anaom. 1979 eogo* 2-231 and 2-333  MOCA OwrmcM Comoony toMUna Ceioraao 90S37
          . «--—	, ^^ Q^, ^ „„ yn^d Juit, tiMimMBonji Proionon Agoner.' tKpiaitiani H n* faaanai Eanon ol SiMfO*n minoea *r «, f.
                                                            	  a« NMor an* WKMOBBUV 1491 Eoajon. Tha uaumninc '**ciari • tonnxiad n a an 01 '00 : 02 >o
                                                            110A lor  a^anaon.  Uootod SlOi tor 91*  rnonu* uonimiainc  groeoov* or uwnao HOC tor in*

                                                                  97-70  1977.
                                                                      A-5

-------

-------
                     APPENDIX B
RECOMMENDED METHODS FOR ORGANIC COMPOUNDS IN EFFLUENT

-------
                TABLE B-l.  LIST OF  TEST PROCEDURES APPROVED  BY  U.S. EPA
                     FOR NON-PESTICIDE ORGANIC COMPOUNDS  IN EFFLUENT


      Note:   This table is an exact reproduction of Table 1C (U.S. EPA 1984),
                                                               IP* i
                                                            QC
                                                                     00 US
                                                                                     OBUT
14,
il
it
17
11
IfL
a
t*
XL «
31 «
                                                           Ml.
                                                           •01.
M
M.
                                                                       •14. 104
                                                                       •14. 1U4
                                                                       •14. I«M
                                                                       U4.1U4
                                            B-l

-------
TABLE  3-1.   (Continued)
        •M ««f M
        807 M ••*. or
        
-------
           TABLE B-2.  LIST OF  TEST  PROCEDURES APPROVED BY  U.S.  EPA
                             FOR PESTICIDES  IN EFFLUENT
Note:  This table is  an exact reproduction  of  Table ID  (U.S.  EPA 1984)
                                                                     ASTM
                                                                       (BOM I Now 3. 0 7. NOW 4. 0. X

                                                                            NOWl. aU.NoWt0.SM.
                                                                            NOWl 0.M NOWC.0 SU
                                                                            NOW! 0 U.NOW* 0 SM
                                                                            NOW! 0 13. NOW* 0 SM
                                                                            Now 1. 0 IS. Now • p SSI
                                                                            NOW 3. 0 10V NOW • 0 SM
                                                                            NOWl 0 7
                                                                                  M NOWl 0 SM
                                                                                  X NOW • 0 S7J
                                                                                  7

                                                                                  104 NOW • 0 5*4
                                                                                  M» NOW 4 0 3S
                                                                                  7 NOW 4 0 30

                                                                                  7 NOW 4 0 X

                                                                                  7
                                                                                  IS NOW 6 0 SSI
                                                                                  « NOW I 0 SSI
                                                                                  M NOW 4 0 X NOW t 3

                                                                                  us
                                                                                  X. NOW • 0 S71
                                                                                  7

                                                                                  7 NOW* 0 X

                                                                                  » NOW • P S71
                                                                                    NOW* P SSI
                                                                                   104 NOW • 0 SM
                                                                                   7
                                                                                   7 NOW 4 p X
                                                                             Now«0X:Naw«.0S73
                                                                             NOW3.0 104, NOW 10 SM
                                                                             NOW 1. 0 10* NOW •. p SM
                                                                             NOW 3 0. 7 NOW 4 p X

                                                                             NOW 1. 0.  7 NOW 4  0. X NOW 6 P
                                                                              sn.
                                            B-3

-------
           TABLE  B-2.   (Continued)
                            »«yu
                                                                                                ASTM
u
                                                                                        MM
                                                                                                     MOW 4 e  X NOW • 0 SM
                                                                                                     Mow 3 0  104 NOW « o W
                                                                                                     MOW3P2SNoW4BXNeUll

                                                                                                     NOW 3 0  04 NOM • V SM
                                                                                                     NOW I. e  i NOW 4 » x
                                                                                                    'NOW! 0  §4 NOW* 0 SM
                                                                                                    'MOW3 0  7
                                                                                                    i MOW 3. 0  104 Now • 0 SS4
                                                                                                    i NOW 3 0  104 NOW I 0 S44
                                                                                                    I NOW 3 0  104 NOW 6 0 S*4
                                                                                                    INOW3 0.M NOW 4 e X
                                                                                                    ' Noli 3
U. PtMnim
M. r i» i •iar
• «• •kBMMBIMVkBIO]
Si. P nciim 	 ,
U ftuowrp 	 __— _^^____
• 1 fin «••!.
M. tim»w 	
fl»*HP 	
M, H.S.T
£ f^Urr^11


fiC
oc
ee
TIC
ne j
Tie
ae
ae
ne
ae
ae
ae
ae.
ae

	





M

NOB
	


— H
MM
MM

i
j Tjua*
	 - -
	
	

OHM

NOW 10 t

NOW 3 J U NOWI 0 SM
NOW 3. 0. U. Now • 0 SM
Now 1 0 U Now •. 0 SM
MM 1 0 104. NOW •. 0 SM
NOW 1 0 M. New • 0 SM
MOW10U. NOW(»1M
M0H 1 0. 104 NOW C 0 M4
MOM Ipo. Mowt. p SM
New 3.0 r
NOM 1 0 104 MOW • 0 St4
Maw 1 0. MI Mew 4. p Jt
MOW 10. Ill
Mew 10 O- MOW* p SM
MOW 10. t M0M4 p 30
MOM 3. p. r

•PMadM on KM « Hi MM oteaim
•Th» M w«t o» iiiirum eot «na to or*
                                                                              nwr bf taw*
                                                                                             TIBW c.
 ! ot BW reiaor laamioi POM      _	
        > I* AOOYM el Oojnc POM***." o( *•> *M 131. Tho i
ill 'pponiti •. "Oitinui 0na ProoMvoier w» Oounvwaan et *w
                                                                                                     Uxnaa DOWCBBI utw  91
                                                                                                                          H'l
       M n««tod oiioi.«un tni IUOU *ar two* wo)
                                                                                          Seek S. OMBWI A) iiiru
                                                                                                             an • rw er«M»M nwa«a
                                                                                                ana tmum 10% • «§ wnowo vwrtM «w
                                                               B-4

-------