NBSIR 73-417
EPA-550/9-73-007
                   Evaluation  of
        Commercial  Integrating-Type
            Noise  Exposure  Meters
                      William A. Leasure, Jr.
                        Ronald L. Fisher
                        Marilyn A, Cadoff

                  NATIONAL BUREAU OF STANDARDS
                         December 1973


                       Joint EPA/NBS Study
               Approved for public release; distribution unlimited
                     Applied Acoustics Section
                        Mechanics Division
                     Institute for Basic Standards
                     National Bureau of Standards
                      Washington, D. C. 20234

-------
NBSIR 73-417
EPA-550/9-73-007

EVALUATION  OF COMMERCIAL  INTEGRATING-
TYPE  NOISE EXPOSURE  METERS
William A. Leasure, Jr., Ronald L. Fisher, and Marilyn A. Cadoff
Applied Acoustics Section
Mechanics Division
Institute for Basic Standards
National Bureau of Standards
Washington, D. C. 20234
December 1973

Final Report
Prepared for
Office of Noise Abatement and Control
U. S. Environmental Protection Agency
Washington, D. C. 20460
U. S. DEPARTMENT OF COMMERCE. Frederick B. Dent. Secretary

NATIONAL BUREAU OF STANDARDS, R.chard W. Roberts. Director

-------
                                                ABSTRACT

As a result of  the promulgation of  occupational noise exposure regulations by the Federal government,
there are a number of commercial  noise  exposure meters on  the market today that provide a measure of
noise Integrated  (with appropriate  weighting) over a time  Interval.  This report presents Che results of
an evaluation of  such Instruments by  the National Bureau of Standards (under the sponsorship of the U. S.
Environmental Protection Agency)  as to  their usefulness in monitoring compliance with occupational
noise regulations as well as  their  applicability as instruments for use in achieving the broader goals
of the EPA.  Tests were designed  and  conducted to evaluate microphone and system response to sound of
random incidence, frequency response, crest factor capability, accuracy of the exchange rate circuitry,
performance of  the noise exposure meter as a function of temperature, and the dependence of the device on
battery voltage.  The rationale of the test procedures utilized to evaluate overall system as well as
specific performance attributes,  details of the measurement techniques, and results obtained are discussed.

Key words:  Acoustics, (sound); dosimeter; environmental acoustics; instrumentation; noise exposure;
            noise exposure meters.
                                                    ii

-------
                                           TABLE OF CONTENTS


1.  Introduction 	     1

2.  Functional Operation of Noise Exposure Meters  	     1

3.  Description of Tests 	     3

    3.1.  Acoustical Tests 	     5

          a.  Microphone Calibration 	     6
          b.  System Response  	     6

    3.2.  Electrical Tests 	     6

          a.  Frequency Response 	     6
          b.  Crest Factor Capability  	     7
          c.  Exchange Rate	     8
          d.  Temperature Response 	     8
          e.  Battery Voltage  	     8

    3.3.  General Observations 	     8

4.  Results of Test	     8

    4.1.  Model A	     8

          4.1.1.  Acoustical Tests 	     8

                  a.  Microphone Calibration 	     8
                  b.  System Response	     8

          4.1.2.  Electrical Tests 	     9

                  a.  Frequency Response 	     9
                  b.  Crest Factor Capability  	    10
                  c.  Exchange Rate	    LO
                  d.  Temperature Range  	    10
                  e.  Battery Voltage	    10

          4.1.3.  General Observations 	    11

    4.2.  Model B	    12

          4.2.1.  Acoustical Tests 	    12

                  a.  Microphone Calibration 	    12
                  b.  System Response	    12

          4.2.2.  Electrical Tests 	    12

                  a.  Frequency Response 	    12
                  b.  Crest Factor Capability  	    12
                  c.  Exchange Rate	    14
                  d.  Temperature Range  	    15
                  e.  Battery Voltage	    15

          4.2.3.  General Observations . i.	    15

    4.3.   Madel C	    15

          4.3.1.  Acoustical Tests 	    15

                  a.  Microphone Calibration 	    15
                  b.  System Response	    15

          4.3.2.  Electrical Tests 	    16

                  a.  Frequency Response 	    16
                  b.  Crest Factor Capability  	    16

                                                   ill

-------
                  c.   Exchange  Rate	   16
                  d.   Temperature  Range	   16
                  e.   Battery Voltage	   16

          4.3.3.   General  Observations  	 ....   16

    4.4.   Model D	   18

          4.4.1.   Acoustical Tests 	   18

                  a.   Microphone Calibration  	   18
                  b.   System Response	   18

          4.4.2.   Electrical Tests 	   18

                  a.   Frequency Response  	   18
                  b.   Crest Factor Capability   	   18
                  c.   Exchange  Rate	   21
                  d.   Temperature  Range   	   21
                  e.   Battery Voltage	   21

          4.4.3.   General  Observations  	   21

    4.5.   Model E	   21

          4.5.1.   Acoustical Tests 	   21

                  a.   Microphone Calibration	   21
                  b.   System Response	   22

          4.5.2.   Electrical Tests 	   23

                  a.   Frequency Response  	   23
                  b.   Crest Factor Capability   	   23
                  c.   Exchange  Rate	   23
                  d.   Temperature  Range	   23
                  e.   Battery Voltage	   24

          4.5.3.   General  Observations  	   25

    4.6.   Model F	   25

          4.6.1.   Acoustical Tests 	   25

          4.6.2.   Electrical Tests 	   25

          4.6.3.   General  Observations  	   25

    4.7.   Model G	   26

          4.7.1.   Acoustical Tests 	   26

          4.7.2.   Electrical Tests 	   26

          4.7.3.   General  Observations  	   26

    4.8.   Model H	   26

          4.8.1.   Acoustical Tests 	   26

          4.8.2.   Electrical Tests 	   27

          4.8.3.   General  Observations  	   27

5.  Summary and Conclusions	   28

6.  Bibliography	   29

-------
                     EVALUATION OF COMMERCIAL INTEGRATING-TYPE NOISE EXPOSURE METERS
                                                   by

                    William A. Leasure, Jr., Ronald L. Fisher, and Marilyn A. Cadoff


                                            1.   INTRODUCTION


      As part of its present and anticipated responsibility for monitoring and controlling noise, the
Environmental Protection Agency, Office of Noise Abatement and Control, has a need to develop and dis-
seminate technical information on the levels and durations of noise to which individuals are directly
exposed.

      Present Federal regulations, promulgated under the authority of the Occupational Safety and Health
Act and the Coal Mine Safety Act, set definite limits on the noise exposure of workers during their 8-
hour working day.  The intent of these regulations is to reduce the risk of permanent noise-induced
hearing damage.

      In order to comply with these regulations, it is necessary to determine an individual's noise
exposure for work environments that have A-weighted-noise levels of 90 dB or higher.  The conventional
method of determining an individual's noise exposure is to use a sound level meter in conjunction with
a detailed time-and-motion study and then calculate the cumulative noise exposure.

      Whether an observer armed with a stop watch and a sound level  meter can accurately characterize
the noise "dose" to which a worker has been exposed is a debatable question.  For constant noise sources
and fixed operator locations, no problem exists.  For the roving worker, however, and the worker who
functions in a fluctuating noise environment, it is difficult to observe their noise exposure and to
compute an accurate daily noise exposure.  This measurement approach is expensive and time-consuming and
is generally inaccurate due to the approximations in the time-and-motion study that practicality dictates.

      The promulgation of occupational noise exposure regulations by the Federal government has resulted
in a proliferation on the market of new sound level meters and, to a lesser extent, noise exposure meters,
or dosimeters, which provide a measure of noise level integrated, with appropriate weighting, over a time
interval.  Because of the motivation for production of these devices, most of them now being manufactured
cover only the range from 90 to 115 dB.

      At present there are no standard performance specifications for such integrating noise exposure
meters.  For this reason, the National Bureau of Standards, under the sponsorship of the Environmental
Protection Agency, conducted a research program to evaluate some available existing noise exposure
meters -- both acoustically and electrically -- as to their usefulness in monitoring compliance with the
OSHA noise exposure regulations and/or carrying out the broader goals of EPA, including determination of
the average individual daily noise exposure of persons in different living patterns.

                            2.  FUNCTIONAL OPERATION OF NOISE EXPOSURE METERS

      Section 1910.95, Occupational Noise Exposure, of the U. S. Department of Labor Occupational Safety
and Health Standards (Federal Register, Part II, Vol. 37, No. 202, October 18, 1972) includes the
following:

      "(a)  Protection against the effects of noise exposure shall be provided when, the sound
      levels exceed those shown in Table G-16 when measured on the A scale of a standard
      sound level meter at slow response 	

      (b)  (1)  When employees are subjected to sound exceeding those listed in Table G-16,
      feasible administrative or engineering controls shall be utilized.  If such controls
      fail to reduce sound levels within the levels of Table G-16, personal protective
      equipment shall be provided and used to reduce sound levels within the levels of
      the table.

      (2)  If the variations in noise level involve maxima at intervals of 1 second  or less,
      it is to be considered continuous.

      (3)  In all cases where the sound levels exceed the values shown herein, a continuing,
      effective hearing conservation program shall be administered.

-------
                                 TABLE G-16  PERMISSIBLE  NOISE  EXPOSURES1


                                                                  Sound level dBA slow
                  Duration per day,  hours                          response

                            8	      90
                            6	      92
                            4	      95
                            3	      97
                            2	     100
                            1 1/2	     102
                            1	     105
                            1/2	     110
                            1/4 or less	     115

      When  Che daily  noise exposure is  composed  of  two  or more periods of noise exposure of
     different levels,  their combined effect  should  be  considered, rather than the individual
     effect of each.  If  the sum of the following fractions:  C1/T1 + C2/T2 + 	 Cn/Tn exceeds
     unity, then,  the mixed exposure should be considered to  exceed the limit value.  Cn
     indicates the  total  time of exposure at  a specified noise level, and Tn indicates the
     total  time  of  exposure permitted at  that level.

     Exposure to impulsive or impact noise  should not exceed  140 dB peak sound pressure
     level."

     The permitted  durations, T, shown  in the above  table can be described by the formula
                                            2/5        •                       


where t is expressed  in hours  and  L    is  the equivalent noise level, expressed in decibels re 20u-PA.

     In general,  the  equivalent  noise  level of  a  time-varying signal of duration T is

                                            r   fT               1n/3
                                            111           3/n
                               Leq  =  10 Iog10b/  (10L/1°)     dtj         ,            (2)



where L is the  time-varying  sound  level and n defines  an  exchange rate between noise level and time.   For
instruments using energy  equivalence,  n = 3, corresponding to a rate of 3.01 decibels per doubling of
time.  Present  U, S.  hearing conservation regulations  (Table G-16 and eq. (1)) use an exchange rate of
n = 4.98, corresponding to a rate  of 5 decibels per doubling of time-

     The equivalent duration of  time-varying signal of actual duration T is

                                                  ,3/n
                                 ,  f
                                  J o
                             fceq  - /   I10     r     I    dc-
where L  is the  rating  sound  level.—   For OSHA  regulations, Lr = 90 dB.
     The percent of allowable  noise exposure  is
                                          ,T
                                       i  l
                           PE  =  100
                                       c,-;
                                          o
                                          ,T   ,.          ->  3/n
                                       1  f    f  (L-Lr)/10|
                                     -J     „        J     ,..
~~ The relation  between eqs.  (2)  and  (5)  is  seen by  observing  that (10L^10)  " = (10L/L0)3
      relation between
     .301^75 „  2L/5  t

-------
 where tr is the rating duration.   Substituting the OSHA rating duration of 8 hr,  the  rating sound  level
 of 90 dB, and n = 4.98, this becomes
                                                           .602
                               PE = 12.5 /   |10V~ '-""I       dt
                                               (L-90)/5
                                              2              dt
                                                                                    (5)
 Thus if L equals 90 dB,  one  will  acquire  12. 5* of  the  allowable  noise  exposure  during  each  hour  exposed
 and  the total allowable  exposure  in  8  hours.   If L equals  95 dB,  25% is  acquired  each  hour  so  that  only
 four hours of exposure are permitted.

      All of the U.  S.-made personal  Integra ting- type noise exposure  meters  investigated  purported to
 measure the quantity defined in eq.  (5).   Because  of the wording of  present  Federal  regulations  the
 devices intentionally do not include levels  below  90 dB in the integration.   In addition to measuring
 percentage of allowable  noise exposure, some of the devices provide  a  means  of  indicating whether or not
 the  sound level exceeded 115 dB during the measurement interval.

      There has been some ambiguity as  to  the  interpretation of Table G-16 in the  OSHA  regulations.  One
 stationary (non-wearable) integrating  noise  exposure meter did not follow eq. (5) but, rather, followed
 the  steps in Table  G-16.  That is, the percent of  allowable exposure was computed from


                                   PE  = i nn
where  Cn  indicates  the  total  time of exposure  in  a specified range of noise  level and Tn  indicates  the
total  time  of  exposure  permitted at  that  level.   Thus Tl = 8 hr corresponds  to  levels in  the  range  90-
92 dB, T2 = 6  hr corresponds  to 92-95 dB, etc.

     Some devices did not measure noise exposure.  The simplest of these caused a warning  light  to  be
turned on when the  noise level exceeded a particular value.  One device measured the total time  that
a particular noise  level was  exceeded.  One device, of European manufacture, did not provide  Information
compatible  with OSHA regulations.  Such devices are not included in this report.

     Basically, a noise exposure meter consists of two parts-  a sound level metering section and an
integrating section.  A block diagram showing  the principal of operation for a  typical dosimeter is
shown  in  Figure 1.  A microphone (most devices utilize an omnidirectional  ceramic microphone) senses the
sound pressure  and  the output is fed into an A-weighting filter which appropriately attenuates the
signal.   The signal is  then detected and averaged to provide an output hopefully equivalent to therms
A-weighted  slow response value that would be read on a sound level meter.

     The  output of  the sound level metering section is then fed into an integrator section which performs
the integration indicated in eq. (5).  The output of the exponent circuit (a voltage) is typically
converted to a  frequency (pulses),  or in the case of the devices that utilize an electrochemical
memory cell, to a current.   The pulse count is accumulated, then monitored and when a given percent of
the allowable  exposure is reached -- for example, one-tenth of one percent -- a signal is sent to the
counter and  the readout display registers one-tenth percent.

     In addition, most devices have a detector which monitors the signal for A-weighted noise In excess
of 115 dB.  If such a noise is detected, an electronic latch Is tripped.   The latch is attached to an
indicating  light and by closing the circuit with a test button, its status can be checked.

                                       3.   DESCRIPTION OF TESTS

     The primary goal of this program was the evaluation of commercial noise exposure meters as to  their
applicability as instruments for use in achieving the broader goals of EPA rather than their usefulness
in monitoring compliance with occupational noise regulations.  Therefore, the study included  testing
of specific performance attributes in addition to overall systems tests.  Well-defined electrical and
acoustical  signals were provided to each device and Che response of the instrument was compared with the
known input.  It was felt that the following factors required attention:

-------
                    -i OVER 115 dB
                    J COMPARATOR
                      OVER IISdB
                      LIGHT
                                                MICROPHONE
                                                A-WEIGHTING NETWORK
                                                SQUARING  CIRCUIT
                                                AVERAGING CIRCUIT
-]  OVER 90 dB
J  COMPARATOR
   EXPONENT CIRCUIT
   (EXCHANGE RATE)

   INTEGRATION /DISPLAY
             T2(L-90)/6dt
       Figure 1.  Block diagram showing the principle of operation for a typical noise exposure meter.


Acoustical Evaluation

     - microphone response  to sound of  random incidence

     - errors or uncertainties due to reflection,  diffraction,  and absorption effects arising from use
       conditions

     - overall system response to sound of random incidence

Electrical Evaluation

     - frequency response

     - detector characteristics (i.e.,  how true is rms response?)

     - dynamic response for time varying signals

     - dynamic range (including internal noise and distortion)

     - nature and accuracy  of time integration

Overall Evaluation

     - appropriateness of quantity measured

     - convenience of use

     - ease  and accuracy of calibration

     - sensitivity to environment

     - durability

-------
     It  would have been prohibitively expensive and time consuming to conduct  detailed calibrations and
physical tests to evaluate  all of the quantities listed above.  Therefore,  the manufacturer's  instruc-
tion manual and wiring diagrams were carefully studied to ensure  that the operation of the device was
fully understood and that  test signals were applied and sampled at appropriate locations.  On  this
basis, carefully selected'tests were carried out to yield the most important information regarding  the
performance of each noise exposure meter.  The following sections contain detailed descriptions  of  the
various  tests -- both acoustical and electrical -- which were performed.

                                       3.1. Acoustical Tests

     Acoustical tests were  performed in a 425 m  reverberation chamber.  For A-weighted-sound  levels at
and below 100 dB, the input signal was broad-band noise shaped  to be "pink".!/  from 50 kHz to 10  kHz.
Above 100 dB an octave band of pink noise centered at 1 kHz was utilized.

     Microphone and speaker placement as well as the signal generation and data  retrieval system utilized
for acoustical tests are sho«m in Figure 2.


            NOISE     SPECTRUM   THREE
          GENERATOR   SHARER  AMPLIFIERS
                                               TO THREE
                                               SPEAKERS

                                              	1
               REVERBERATION
                 CHAMBER
                                          ©
MICROPHONE
PLACEMENT
DETAIL
                                                                   FROM
                                                                   MICROPHONE
                              TELETYPE COMPUTER  REAL-TIME
                                                       ANALYZER

           Figure 2.   Block diagram of  equipment used for acoustic tests in a  reverberation chamber.
2f  "pink" noise  is white noise  passed through a network which weights  at -3 dB per octave.

-------
 Sound pressure Levels were measured using a one-inch condenser microphone.

      For all tests the vanes in  the chamber were rotated at 5 r.p.m. to promote diffuseness of the sound
 field.

                                       a.  Microphone Calibration

      During microphone calibration the reference microphone was located at position 1 (see Figure 2)
 and noise exposure meter microphones (test microphones) occupied the remaining six positions on the
 circumference of a two-foot diameter circle surrounding the standard microphone.   All microphones were
 located five feet above the floor.

      The noise exposure meter microphones had been removed from the devices.   Care was taken to main-
 tain microphone cable length.  The signal from each test microphone was fed into a preamplifier and
 then into a one-third octave band real-tine analyzer.

      The sound level of the pink noise was adjusted to read 100 dB on the A scale of the real-time
 analyzer.  Other sound levels were utilized to check the linearity of the microphones over the range
 of  interest (90-115 dB) .

      Once the response of the measuring system -- reference microphone, preamplifiers and real-time
 analyzer -- had been set utilizing a pistonphone and the sound field established, the calibration was
 placed under computer control.   The one-third octave band correction values based on electrostatic
 actuator calibration of the reference microphone were stored in the computer.   The computer was pro-
 grammed such that signals from the reference microphone were interrogated 120  seconds (approximately
 6  time constants in the slow-random mode of the real-time analyzer) after the  onset of the sound
 fields.   Addition of the correction values and the response of the reference microphone to the sound
 field resulted in the establishment of the desired one-third octave band sound pressure levels in the
 room.  In sequence each of the test microphones were interrogated in a similar manner.  Using these
 data, the computer calculated the response of each test microphone.

                                           b.   System Response

      For system response testing, the microphone of each noise exposure meter  was positioned on the
 circumference of a two-foot diameter circle which surrounded the reference microphone.  All microphones
 were located five feet above the floor.   The response of the measuring system  was set with a pistonphone
 (124 dB at 250 Hz), the desired sound field was established, and the sound field  was turned off.  Once
 all  of the noise exposure meters were turned on and zeroed, the sound field was turned on, at the
 established level, for a time equivalent to 75% of the permissible noise exposure under OSHA regulations.
 This was chosen rather than 100% since one device would not indicate overranging  above 100%.  Upon
 completion of the appropriate time period, as monitored by a stop watch, the sound field was turned off
 and  the noise exposure meter readings were recorded.   This procedure was utilized for sound fields of
 92,  95,  100,  105, 110 and 115 dB.

                                          3.2.  Electrical Tests

      In this section, descriptions are given of the various tests which were carried out to determine
 specific attributes of device performance.

      For these  tests an electrical signal was inserted in series with the microphone.   A signal
 generator with  a low output impedance,  compared to the impedance of the microphone,  was used to ensure
 that the total  input impedance  to the  electronics was essentially the same as  in  normal usage.

      In principle, all of these tests  could be performed using the readout device provided with the noise
 exposure meter.   However,  because of  the large quantity of data  being acquired this  would have  been
 prohibitively time-consuming.   Accordingly, for one of the tests (see 3.2.a.) .voltage measurements were
 made at  an internal point in the instrument.   For other tests,  direct measurements were made of the
 period of the pulse train which advanced the  counter  (mechanical or electronic)  in the noise exposure
 meter.   In many cases,  a  fairly high  frequency pulse  train was  available (with a  dividing circuit to
 produce  a lower frequency signal to actuate the counter)  so that the time to obtain  a datum point was
 reduced  from possibly hours to  the few seconds needed for  the  system to stabilize after changing the
 input signal.

                                         a.  Frequency Response

      The frequency response was measured by injecting a sine wave voltage in series  with the microphone
 and  then measuring the voltage  at the  output  of the A-weighting  network while  the frequency of  the test
 signal was  swept  over the  frequency range from 20 Hz  to 20 kHz.   The frequency response was recorded
directly on a graphic level recorder which had its  paper speed  synchronized with  the oscillator sweep
 speed.

-------
     The frequency  response of the electronics was combined with that oE  the microphone (see  3.1.a.),
to obtain the overall frequency response curves shown in Section 4.

                                     b.  Crest Factor Capability

     A pulsed sine  wave was presented to the  noise exposure meter via the voltage insertion technique to
obtain a measurement of the crest factor handling capability.   The crest  factor of a signal is defined
as the ratio of  peak signal value to rms (root-mean-square) value.  It is important to consider because
many industrial  noises have a high crest factor.

     The pulsed  sine wave consisted of a 1 kHz sine wave gated  by a  pulse train with a frequency of 100
Hz (period = 10.0 ms) and an adjustable pulse duration.   By changing the pulse duration from  full-on
to a very short  pulse (12.5 ms),  the crest factor was increased from 1.414 (sine wave) to 4.0.  As the
pulse duration was  decreased, the pulse amplitude was increased in order  to maintain a constant rms
voltage.  The rms voltage levels  applied were those corresponding to the  114 dB and 95 dB levels for the
individual noise exposure meters.

     The equipment  setup  is shown in Figure  3.  The gating of  the sine wave was accomplished using an
analog multiplier.  A precision ac voltmeter  (accurate for signals having a crest factor up to 10)
was used to measure the rms value of the pulsed sine wave.  An  oscilloscope was used to msasure peak
values.  The response of the noise exposure meter was determined  either by measuring the period of the
pulse train which actuated the counter or by  using the actual  readout from the noise exposure meter.
                         SINE WAVE
                         OSCILLATOR
                   TRUE RMS
                   VOLTMETER
                         I
                         *
     PULSE
   GENERATOR
                                                     ANALOG
                                                     MULTIPLIER
                                                                o
                        OSCILLOSCOPE
STEPPED
ATTENUATOR
                            MICROPHONE
                                                        IN
          OUT
                                                                                 DIGITAL
                                                                                 COUNTER
                                                   NOISE EXPOSURE
                                                        METER
                Figure 3.  Block diagram of  the equipment used  for crest factor measurements.

-------
                                            c.  Exchange Race

      Using the voltage Insertion  technique, a  1 kHz sine wave was injected in series with the microphone
and the response of the noise exposure meter observed  (either directly or by observing the internal pulse
train) as the voltage was varied over a  range corresponding approximately to a sound level of 90 to 115
dB.

      The data — hours for 100Z exposure vs. the  input voltage level (i.e., vs. the logarithm of voltage)
— were plotted and a straight  line was  drawn through  the central portion (main trend) of the data.  The
voltage corresponding to an allowable exposure  time of 0.25 hr was then read from this curve.  Using
this voltage in an expression analogous  to  eqs. (4) and  (5) in Section 2, calculations were made of the
percent of allowable noise exposure which the device would be expected to read for each test voltage
provided it were functioning perfectly.  The observed  readings (actual data points) were then ratioed to
the calculated expected readings.  In effect, this procedure compares the response of the device at any
test voltage to the response which would be expected if  the device (1) were functioning perfectly at a
voltage corresponding to a noise level of 115 dB and (2) had exactly the correct exchange rate of 5
decibels for a doubling of exposure  time.

                                         d.  Temperature  Response

      Since personal noise exposure meters  may  be  used in occupations where temperature extremes occur,
limited measurements were made  of  performance as a function of temperature.

      A 1 kHz signal was injected  in  series with the microphone and the output of the noise exposure
meter measured with the device  at  room  temperature (24°C), in a small oven at 45+l°C, and in a refrigera-
tor at 5+l°C.  The instruments  were  allowed to  reach thermal equilibrium before data were taken.

                                            e.   Battery Voltage

      Personal integrating noise exposure meters are battery operated and during normal usage the battery
voltage will decrease.  To determine  the dependence of the dosimeter on battery voltage, an adjustable
dc power supply was substituted for  the  Internal battery.  The voltage was adjusted over a wide range
appropriate for the battery  type  (in a  specific voltage  range given by the manufacturer) while injecting
a 1 kHz signal in series with  the  microphone,   Thus  the  performance of the instrument was determined as
a function of supply voltage.   Most  of  the  devices had a voltage  regulator and a battery check feature
to indicate when the battery voltage was too  low for proper operation.  When  these  features were function-
ing properly, the tests indicated  that  low  battery voltage would  not result in erroneous measurements
unless the battery was sufficiently  discharged  to  deactivate  the  battery check indicator.  For one
instrument, however, readings  of noise  exposure were found  to be  significantly in error before the
battery check indicator revealed a problem.  For  this  instrument  the performance is reported at the
voltage which deactivated  the  battery check indicator.

                                         3.3.  General Observations

      In addition to  the  specific  acoustical  and  electrical  tests described above,  observations were made
relative to  the convenience  of use and  to  factors  which  could  lead to maintenance or  operational difficul-
ties.  These observations  are  listed in Section 4  for  each Instrument  evaluated.

                                           4.   RESULTS  OF TEST

      In  this  Section,  results are given for the tests descrioed  in Section  3.   The test  samples included
the  following  commercial  noise exposure meters:  Columbia models  101 and  104,  Dupont,  General  Radio models
1934  and  1944,  3M,  Quest  M-6,  and Tracoustics.   Unless otherwise  Indicated,  two samples of  each model
were  tested. All  of the  models tested were purchased between March and  May,  1972;  therefore,  present
models may not  be identical to those teste  due to possible modifications  by  the manufacturer.

                                       Personal (Wearable) Instruments

                                               4.X. Model A

                                          4.1.1- Acoustical Tests

                                         a.  Microphone Calibration

       The relative response of the microphone based on 0 dB response at 1 kHz, is shown as a function of
 frequency in Figure 4.1-1.  Tests over  the range  of A-weighted-sound levels from 90 to 115 dB indicated
 no problems with linearity.

                                            b.  System Response

       The overall performance  of  the noise exposure meter, when placed in a random, diffuse sound  field
 as described in Section 3.1.b.,is shown in Figure 4.1-2.  The response was measured relative to a
                                                     8

-------
calibrated  condenser microphone and measurement system.  The cross-hatched region indicates  the estimated
uncertainty (95 percent confidence limits) in the  level of the sound field in which the noise  exposure
meter was  tested.
                   20
     10
              O
              •o
              uT
              CO
              Q.
              CO
              Ul
              <
              d  -'o
              K
                  -20
                                        I
                                  1       I
         -   MODEL  A
                                      o  SAMPLE I
                                      A  SAMPLE 2
                           I
                          I
I
                                0.1
                                         1.0
                              FREQUENCY, KHz

Figure 4.1-1. Relative Frequency response of microphone.
                            10.0
RELATIVE RESPONSE, PERCENT
g g 1 § I
1 1
— MODEL A
o
0
1
O
1
t
1
—
\\X\\\\\\\\\\\\\\\\\\\\\\
\\\\v\\v\\\\\\\\\\\\\\\\\
~ 0 SAMPLE 1
A SAMPLE 2
I |
1
0
A
1
o
A
1
8
A
i
—
                           90       95      100       105     110       115
                               A-WEIGHTED SOUND LEVEL,dB reZOftPo

             Figure 4.1-2. Response of noise exposure meter, relative  to expected response
                          when placed in a field of pink noise (see  text) at A-weighted-
                          sound levels from 92  to  115 dB.

                                     4.1.2.  Electrical Tests

                                     a.   Frequency Response

     The  combined frequency response of the microphone, the input amplifier, and the A-weighting network,
as measured at the output of  the A-weighting network, is  shown in Figure 4.1-3.

-------
                20
                10
           ffi
           •o
           uf
           tn
              - IO
           >  —20
           <
           UJ
              -30
              -40
              -5O
                        MOD
                     INSERT
                                             o  SAMPLE  I
                                             A  SAMPLE  2
                                               \
                                                                         I    _
                                                                          I   -
                            O.I
                            1.0
                 FREQUENCY, kHz
       10.0
         Figure 4.1-3.
Relative combined frequency response of microphone plus
electronics.  The dashed curves indicate the allowable response
level limits for a Type 2 sound level meter as specified in
American National Standard Specifications for Sound Level Meters
SI.4-1971.
                                     b.  Crest Factor Capability

     The response of the noise exposure meter,  normalized to 100% for a crest factor of  1.414 (sine wave),
is shown in Figure 4.1-4. as a function of the  crest factor (ratio of peak voltage to rms voltage)  of  the'
test signal.

                                         c.   Exchange Rate

     The response of the noise exposure meter,  normalized to 100% at a duration of 0.25  hours, relative
to the response of an instrument with an exchange rate of exactly 5 decibels per doubling of time Is
shown In Figure 4.1-5.  The important thing to  consider here is whether or not the data  points define  a
flat curve  over a range of at least 25 dB.  Deviations from flatness at one end of the 25 dB range  shown
could be compensated for, by gain adjustments,  if the curve is flat over a total of at least 25 dB.

                                       d.  Temperature Range

     The response of the noise exposure meter,  at high and low temperatures, relative to 100% response at
24°C, was found to be:
                      5°C
                     45°C
                Sample 1
                106%
                106%
                e.  Battery Voltage
Sample
  94%
 113%
     The  response of the noise exposure meter, relative to  100% response  for a full-charged battery, was
75% for Sample 1 and 91% for  Sample 2 at the voltage at which the battery check indicator showed  battery
failure.
                                                10

-------
ERRATUM*
NBSIR 73-417,  EPA-550/9-73-007

Replace Figure 4.1-3,  page 10, with figure below:
   +20



   +10
    -10
    20
   -30
    40
    -50
 MODEL A
I     I
I
                              e SAMPLE 1

                              A SAMPLE 2
              0.1
                    1.0

             FREQUENCY. kHz
                   10.0
Replace Figure  4.2-3,  page 13,  with figure below:

     20
     10
  8
  uT
    -10
    -30
   -40
   -50
          I

          MODEL B
                 I     I     I
                 I     I
       -    /r/
I    I     I
                          I    I     I
              0.1               1.0

                       FREQUENCY. kHz
                                     10-0

-------

120
t-
8 100
tf
*
Ul
z 8O
2
CO
Ul
tc
u, 60
p
-i
c 40

2O
1 1 1 1 1
~~ MODEL A ~~
— —
A
u
_ OIISdB —
A A 95dB
~ O ~~
— A —

— A —
_ 0 _

— ' —
0
1 1 1 1 1
12345
                                            CREST FACTOR


                      Figure 4.1-4.   Response  of  noise  exposure meter, normalized to 100% for

                                     a  sine  wave  input,  as a function of crest factor.
                140
              Ul
              u
              tc
              CO


              o


              CO
              Ul
              IE


              Ul
              Ul
              tc
                 120
100
                 60
1
~ 1
_

MODEL A
io
A 0^0o.0rt00o
° v fft> o° A O A
A
A
^
• A ^ ^^^^dD
0 SAMPLE 1
A SAMPLE 2
—
oA A
A ~~
o «
o
o
              Figure  4.1-5. Relative response of noise exposure meter to an electrical  signal

                           covering a range of approximately 25 dB.  See text  for method of

                           normalization.


                                      4.1.3- General Observations



      a.   This instrument did not appear to have a voltage regulator.



      b.   The microphone shield lead was soldered to the case hinge and  relied on electrical conduction

from one  side of  the  hinge to the other.
                                                  11

-------
                                     Personal (Wearable) Instruments

                                              4.2. Model B

      Two samples were purchased but tests are reported for only one since the other was  found to be
defective.

                                         4.2.1. Acoustical Tests

                                        a.  Microphone Calibration

      The relative response of  the microphone,based on 0 dB response at 1 kHz, is shown as a  function of
frequency in Figure 4.2-1.  Tests over the range of A-weighted-sound levels from 90 to 115 dB indicated
no problems with linearity.

                                           b.  System Response

      The overall performance of  the noise exposure meter, when placed In a random, diffuse sound field
as described in Section 3.1.b.,is shown in Figure 4.2-2.  The response was measured relative  to  a cali-
brated condenser microphone and measurement system.  The cross-hatched region indicates the estimated
uncertainty (95 percent confidence limits) In the level of the sound field in which the noise exposure
meter was tested.
                 ul
                 CO
                 o.
                 CO
                 UJ
                 QL

                 UJ
                      20
                       10
                 UJ
                 tc
                    - 10
                    -ZO
    i      i       r
h-  MODEL  B
                                                      I       I
                                                                             I      I
                                    •• • •
                                    0.1
                                                         10.0
                                                            1.0

                                                 FREQUENCY, kHz

                 Figure 4.2-1. Relative frequency response of microphone.

                                          4.2.2.  Electrical Tests

                                          a.   Frequency Response

      The combined  frequency  response of  the  microphone,  the  Input amplifier, and the A-welghting network,
as measured at  the  output  of  the  A-weigb.ti.ng  network, is  shown in Figure 4.2-3,  The  manufacturer of
this instrument has notified  MBS  that they have'modified  the  design to improve the high frequency perfor-
mance.
                                       b.  Crest Factor Capability

     The response of  the noise exposure meter,  normalized to  100% for a crest factor of 1.414 (sine wave),
Is shown in Figure  4.2-4 .as a function of the crest  factor (ratio of peak voltage to rms voltage) of the
test signal.
                                                    12

-------
   140
  UJ

  oe 120
  o 100
  a.
  V)
  UJ
  o:
  UJ
  > 80


  §
  UJ
  (T

    60
       —  MODEL B
              I
     I
I
I
I
I
             90     95     100     105     110     115


                A-WEIGHTED SOUND LEVEL, dB re 20 ftPo


 Figure 4.2-2. Response of noise exposure meter, relative to expected response,

           when placed in a field of pink noise (see text) at A-weighted-sound

           levels from 92 to 115 dB.
     20
     IO
  m
  ^

  uf


  O -10
  Q.
  W
  UJ
  K
    -20
 E
 UJ
    -30
    -40
    -5O
           MODEL B
i     i       i      i     r     i     i

           INCORRECT
Y SEE  INSERT
               O.I
                  i.o
                        10-0
                        FREQUENCY, kHz

Figure 4.2-3. Relative combined frequency response of microphone plus electronics.

          The dashed curves indicate the allowable response level limits for a

          Type 2 sound  level meter as specified in American National Standard

          Specifications for Sound Level Meters, SI.4-1971.
                              13

-------
120
gioo
ui
0.
RESPONSE,
OD
O
UJ 60
£
i
UJ
K 40
20
1 1 1 1 1
_ MODEL B _
~~ 0 ~
A
° A °
_ A
a
~~ "••
~ 0 1 1 5 dB ~~
_ A 95 dB _
1 1 1 1 1
                                         12345
                                               CREST FACTOR

                         Figure 4.2-4.  Response of noise exposure meter  normalized
                                       to 100% for a sine wave input,  as a  function
                                       of crest factor.
                                           c.  Exchange Rate

     The response of the noise exposure meter, normalized to 100% at a duration  of 0.25  hours, relative
to the response of an instrument with an exchange rate of exactly 5 decibels  par doubling of  time is
shown in Figure 4.2-5.  The important thing to consider here is whether  or  not the data  points define
a flat  curve over a range of at least 25 dB.  Deviations from flatness  at  one end of  the 25  dB range
shown could be compensated for, by gain adjustments, if the curve is flat over a total of at  least 25 dB.

I4O
PERCENT
N
O
UJ
U)
° inn
O. IOO
tr
UJ
P 8°
<
_j
u
QC
60

—
™*™
•^
i
—





MODEL B

• • — ». A & A
, . . -• ••-.







:
~^
r
—




          Figure 4.2-5. Relative response at noise exposure meter  to  an electrical
                        signal covering a range of approximately 25 dB.   See text
                        for method of normalization.
                                                   14

-------
                                         d.  Temperature Range

     The response of the noise exposure meter, at high and low temperatures,  relative  to  100% response
at  24°C, was found to be:

                                 5°C        102%
                                45°C         99%

                                          e.  Battery Voltage

     Operation of the battery check indicator appeared satisfactory -- as long as  the  light would go on,
the response of the noise exposure meter did not change.

                                      4.2.3. General Observations

     a.  Turning the switch from "on" to "battery check" and back to "on" was observed to advance the
counter by one count (.1%).

                                    Personal (Wearable) Instruments

                                             4.3. Model C

     Only one sample was tested.

                                        4.3.1. Acoustical Tests

                                      a.  Microphone Calibration

     The relative response of the microphone,based on 0 dB response at 1 kHz, is shown as a  function
of  frequency in Figure 4.3-1.  Tests over the range of A-weighted-sound levels from 90 to 115 dB
indicated no problems with linearity.
          CD
           bl
           o
           Q.
           v>
           UJ
           or
           u
               20
                10
              -10
          uJ
             -20
    I       I
—  MODEL  C
                                                     T
T
T
                                              I	I
                            O.I
                                 1.0
                10.0
                                        FREQUENCY, kHz
         Figure 4.3-1. Relative frequency response of  microphone.

                                          b.   System Response

     The overall performance of the noise exposure meter,  when  placed  in  a  random, diffuse sound field
as described in Section 3.1.b.,is shown in Figure  4.3-2.   The response was  measured relative to a
calibrated condenser microphone and measurement  system.  The cross-hatched  region indicates the estimated
uncertainty (95 percent confidence limits) in the  level of the  sound field  in which the noise exposure
meter was tested.
                                                   15

-------
                  140
               UJ
               u
               id 120
               UJ
               CO
                OL 100
                UJ
                tr.
                UJ
                I  80
                UJ
                Q£
                   60
     I         I
MODEL  C
                                          I
                                          I
                               90      95       100      105      110       115

                                  A-WEIGHTEO SOUND LEVEL, dB re 20/j.Pa

              Figure 4.3-2.  Response of noise exposure mecer, relative Co expected  response,
                            when  placed in a field of pink noise (see  text)  at A-weighted-
                            sound level from 92 to 115 dB.
                                        4.3.2.  Electrical Tests
                                        a.  Frequency response

     The combined frequency response of the microphone,  the  input amplifier, and the A-weighting network,
as measured at the output of the A-weighting network,  is  shown in Figure 4.3-3.


                                      b.  Crest Factor Capability

     The response of the noise exposure meter,  normalized to 100% for  a crest  factor of 1.414 (sine wave),
is shown in Figure 4.3-4. as a function of the crest factor (ratio of peak  voltage to rms voltage) of the
test signal.

                                           c.  Exchange Rate

     The response of the noise exposure meter,  normalized to 100% at a duration of  0.25 hours, relative
to the response of an instrument  with an exchange rate of exactly 5 decibels per doubling of time is
shown in Figure 4.3-5.  The important thing to consider here is  whether or not the  data points define
a flat curve over a range of at least 25 dB.  Deviations from flatness at  one  end of the 25 dB range
shown could be compensated for, by gain adjustments,  if the  curve is flat  over a total of at least
25 dB.

                                         d.  Temperature Range

     The response of the noise exposure meter,  at high and low temperatures, relative to 100% response
at 24°C, was  found to be:
                           5°C
                          45"C

                                          e.  Battery Voltage

     This  instrument did not have a battery voltage indicator.

                                      4.3.3- General Observations

                                         (none)
                                     100%
                                     101%
                                                   16

-------
            20
            IO
        m
        •Q
        u
        M
        I  -.0

        J3
        oe

        >  -20


        u
           -30
           -40
           -50
                  T	1

                    MODEL C
T     T
                                                              I  _
                                                         I	L
                       O.I                  I.O
                                  FREQUENCY, kHz
                                                              10X3
Figure 4.3-3.  Relative combined frequency response of microphone plus
              electronics.  The dashed curves Indicate the allowable
              response level limits for a Type 2 sound level meter  as
              specified  in American National Standard Specifications  for
              Sound Level Meters, SI.4-1971.

I2O

8 ioo
(T
HI
Q.
ISPONSE,
0»
O
ud
(T
u 60

u
o: 40
ZO
1 1 1 1 1
— MODEL C —
_ 0 1 1 5 dB _
A 95 dB
_ 0 _
~~ A ~~
0 _
_ A _
— o —
— —
1 1 1 1 1
                           12345

                                 CREST FACTOR

         Figure 4.3-4. Response of noise exposure meter, normalized
                       to 100% for a sine wave  input as a function
                       of crest factor.
                                      17

-------

I4O
t-
z
Ul
o
£.20
Q. IOO
en
Ul
tr.
Ul
C 80
r~
_J
Ul
OC
60

—
_


'
—


^~

~
1
1 MODEL C
1
1

\ ' 	


<13 Oo 9




—
_

—
. -
— —


^^

^^^
           Figure it.3-5. Relative response of noise exposure meter to an electrical signal
                         covering a range of approximately 25 dB. See text for method of
                         normalization.
                                     Personal (Wearable) Instruments

                                              4.4.  Model D

     Two  pilot-production samples, designated 1 and 2, were tested.   The  overall  (acoustical) system
performance  of  two  production samples, designated  3 and 4,  was  determined.

                                         4.4.1.  Acoustical Tests

                                       a.  Microphone Calibration

     The  relative response of the microphone, based on 0 dB  response  at  1  kHz,  is  shown  as a function of
frequency in Figure 4.4-1.  Tests over the  range of A-weighted-sound levels  from  90  to  115 dB indicated
no problems  with linearity.

                                          b.   System Response

     The  overall performance of the  noise exposure meter, when  placed in  a random, diffuse sound field
as described  in Section 3.1.b.,is shown in  Figure  4.4-2.  The response  was measured  relative to a cali-
brated condenser microphone  and measurement system.   The  cross-hatched  region  indicates the estimated
uncertainty  (95 percent confidence limits)  in the  level of  the  sound field in  which  the noise exposure
meter was  tested.

                                        4.4.2.  Electrical Tests

                                        a.   Frequency Response

     The  combined frequency  response  of the microphone,  the  input amplifier, and  the A-weighting network,
as measured at the  output  of the A-weighting  network,  is  shown  in Figure  4.4-3.

                                       b.  Crest  Factor Capability

     The  response of  the noise  exposure meter,  normalized to 100% for a crest  factor of 1.414 (sine wave),
is shown  in Figure  4.4-4.as  a function of the crest  factor  (ratio of  peak voltage to rms voltage)  of the
test signal.
                                                   18

-------

20
€
ME RESPONSE,
0 0
£
u-IO
ac

-20

"1 T
— MODEL D


frte*^
o SAMPLE 1 S^
A SAMPLE 2
— —
III III II
                 O.I                    1.0

                             FREQUENCY,  kHz

Figure  4.4-1. Relative frequency  response of  microphone.
                                                              10-0
RELATIVE RESPONSE, PERCENT
n 0» Q ro £
o o o o o
-T- n~ -i i i" ~r
MODEL D
l_ A
* A 4 • t
\\\\\\^\\\\\\\
\\ V\»\ V\ V\ V\ \9\ \
A
" o SAMPLE 1
- A SAMPLE 2
• SAMPLE 3
~ A SAMPLE 4
I 1 1 1 1 1
90 95 100 105 110 115
                 A-WEIGHTED SOUND LEVEL, dB re 20


Figure 4.4-2. Response of noise exposure meter, relative to expected response,
             when placed in field of pink noi«e (see text) at A-weighted-sound
             levels from 92 to 115 dB.
                                   19

-------
       2O
        10
      -10
      -30
      -40
      -50
               MODEL 0
                      o  SAMPLE  I
                      A  SAMPLE  2
                   O.I                   I.O

                              FREQUENCY, kHz
                                             10-0
Figure 4.4-3.   Relative combined frequency response of microphone plus
               electronics.  The dashed curves indicate the allowable response
               level limits for a Type 2 sound level meter as specified
               in American National  Standard Specifications for Sound Level
               Meters, SI.4-1971.

120
:, PERCENT
0
O
|ao
CO
UJ
oc.
ui 60
\
£ 40
20
1
MODEL D
A a a a a
u
— —
0 1 15 dB
_ A 95 dB
1 — —
1 1 1 1 1
 Figure 4.4-4.
        12345
              CREST FACTOR
Response of noise exposure meter, normalized to 100%  for a
sine wave  input, as a function of crest factor.
                                  20

-------
                                           c.   Exchange  Rate
     The response of the noise exposure  meter,  normalized  to  100% at a duration of 0.25 hours, relative
to the response of an instrument with an exchange rate of  exactly 5 decibels  per doubling of  time  is
shown in Figure 4.4-5.  The important thing to  consider here  is whether or not the data points define a
flat curve over a range of at least 25 dB.   Deviations from flatness at one end of the 25 dB  range shown
could be compensated for, by gain adjustments,  if the curve is  flat over  a total of  at least  25 dB.

                                         d.  Temperature Range

     The response of the noise exposure  meter,  at high and low  temperatures,  relative  to  100% response
at 24°C, was found to be
                               5«C
                              45°C
Sample 1

  101%
Sample 2
  96%
 102%
                140
             UJ
                120
              O
              (L
              id
              IT
              UJ
              3
              UI
                 80
                 60

_ A

MODEL D
" ° nCwO— & A O O— O
O A A ** " A Mi A A A /
o^
o SAMPLE 1
A SAMPLE 2

& _

           Figure 4.4-5.  Relative response of noise exposure meter to an electrical signal
                         covering a range of approximately 25 dB.  See text for method of
                         normalization.

                                          e.  Battery Voltage

     Operation of the Battery check indicator appeared satisfactory.

                                      4.4.3, General Observations

     a.  It was observed that the unit can be permanently damaged during battery installation if the
battery  terminals are touched with reversed polarity to the connector.

     b.  Difficulty was experienced in connecting the production noise exposure meters to the calibrator
due to mechanical misalignment.

     c.  The normal calibration check was not sufficiently precise (+ 10% of permissible noise exposure).

                                    Personal (Wearable) Instruments

                                             4.5. Model E

                                        4.5.1  Acoustical Tests

                                      a.  Microphone Calibration

     The relative response of the microphone, based on 0 dB response at 1 kHz, is shown as a function of
frequency  in Figure 4.5-1.   Tests over the range of A-weighted-sound levels from 90 to 115 dB indicated
no problems with linearity.
                                                   21

-------
              m
              •o
              UJ*
              en
              Q.
                80
          UJ
          o:
              60
         	1	T

       —    MODEL E
               O  SAMPLE I
               A  SAMPLE 2
                         _L
                        _L
_L
_L
Figure 4.5-2.
                         90       95      100     105       110      115
                           A-WEIGHTED SOUND LEVEL, dB re 20/iPd

                        Response of noise exposure meter,  relative to expected response,
                        when placed in a field of pink noise (see text) at A-weighted-
                        sound levels from 92 to 115 dB.
                                                22

-------
                                       4.5.2.  Electrical Tests

                                       a.   Frequency Response

     The combined  frequency response of Che microphone,  the input amplifier, and the A-weighting network,
as measured  at  the output of the A-welghtlng network, is shown in Figure 4.5-3.   It should be noted that
this unit has a potted electronics module;  therefore, measurements had to be made at the output of the
exchange rate circuit rather than at the output of  the A-weighting network.  For this  reason one could
only check the  frequency response over a 25 dB  range.
                      20
                       IO
                CD
                TJ
                uf
                CO
                £   -10
                a
                cr
                UJ

-------
at 24°C, was found to be:
u uv .
5°C
45°C
I4O
Ul
o 120
Ul
Q.
RESPONSE,
O
O
uj 80
<
Ul
«r 6O

4O
Sample 1 Sample 2
100% 99%
101% 102%
1 1 1 1 1
~~ MODEL E ~~
_ A _
a
A
G o
^— —
_ o —
~~ o I 15 dB ~
_ A 95 dB —
1 1 1 1 1
                                              234
                                             CREST FACTOR
                     Figure 4.5-4.  Response of  noise exposure meter, normalized
                                   to 100%  for  a  sine wave input, as a function
                                   of crest factor.
ONSE, PERCENT
> ro Z
> o O
Q. IUU
tr
S 80
UJ
a 6O
]
— o

MODEL E
A /V» n o
o o
«""*^ rlrt hi
0 SAMPLE 1
A SAMPLE 2
~
_
O.
—
           Figure 4.5-5.   Relative response of noise exposure meter to an electrical  signal
                          covering a  range of approximately 25 dB.  See text  for  method of
                          normalization.
                                            e.  Battery Voltage
     The response of the noise exposure meter, on both  samples, relative to 100%  for  a  fully-

                                                   24

-------
charged battery, was 94% at  the  voltage  at which  the battery check indicator showed battery failure.

                                     4.5.3. General Observations

     a.  The clip for attaching  the microphone  to  the user's clothing easily became detached from the
microphone.

                                   Personal (Wearable) Instruments

                                            4.6. Model F

     These devices were received very  late  in  the  program.  Only the overall acoustical performance and
the exchange rate were evaluated.

                                        4.6.1.  Acoustical Tests

     The overall performance of  the  noise exposure  meter, when placed in a random diffuse sound field
as described in Section 3.1.b.,is shown in Figure 4.6-1.  The response was measured relative to a cali-
brated condenser microphone and  measurement  system.  The cross-hatched region  indicates the estimated
uncertainty (95 percent confidence limits)  in  the level of  the sound field in  which the noise exposure
meter was  tested.
RELATIVE RESPONSE, PERCENT
0> o> O w *
O O O O O
1 1 1
- MODEL F
1 1 1
\\\\\\ \\\\\\\\\\\\\\\\
\\\\\\X\\^\>
o
A
0 SAMPLE 1
A SAMPLE 2
1 1 1

\
6
i i i
^™
\
s
^
;
                              90       95      100      105      110       115
                                A-WEIGHTED SOUND LEVEL,dB re 20 /APa
            Figure 4.6-1. Response of noise exposure meter,  relative  to expected  response when
                          placed in a field of pink noise (see text)  at A-weighted-sound levels
                          from 92 to 115 dB.

                                        4.6.2. Electrical Tests

      The response of  the noise exposure meter, normalized to 100% at a duration of 0.25 hours,  relative
 to the response  of an instrument with an exchange rate of exactly 5 decibels per doubling  of time is
 shown in Figure  5.6-2.  The important thing to consider here is whether or not the data points  define
 a £lat curve  over a range of  at least 25 dB.  Deviations from flatness at one end of the 25 dB  range
 shown could be compensated for, by gain adjustments,  if the curve is flat over a total of  at least
 25 dB.  It should be  noted that sample 2 is a first generation design while  sample 1 is a  later   model.
 The manufacturer of this instrument notified NBS that  they had modified the  design to cover the necessary
 25 dB dynamic range.

                                      4.6.3. General Observations

                                                (none)
                                                  25

-------
140
i
2JI20
uf
CO
0
^k I^X^X
CO
u

-------
        140
        120
    5
    o
    UJ
    (/>

    o
    Q.
    (rt
    UJ
    ac

    UJ
        80
       100
    UJ
    (K
        60
            	1	T

           !-   MODEL G
                      044
O  SAMPLE I

A  SAMPLE 2
                   I
Figure
                   90      95      100      105      110       115

                    A-WEIGHTED SOUND LEVEL, dB re 20 ft Pa


        .7-1.  Response of noise exposure meter, relative to expected response, when

              from 921?oamedt°f Plnk n°1Se (S6e teX° " A-»"8"ted-sound ieve?"
       140
     u
     o
     o:
        120
     V)


     glOO

     UJ
     tc
        80
    UJ
    
-------
                                      5.  SUMMARY AND CONCLUSIONS

      In general, the following conclusions can be drawn as a result of this test  program:

         Microphone response — Most noise exposure  meters utilize well-proven ceramic microphones
                    with characteristic response being relatively flat  from 50 Hz to 5-8 kHz.
                    None of the microphones tested showed any evidence of nonlinearities over
                    the dynamic range of Interest (90-115 dB).

         System response — Since no performance standard exists against which these devices can
                    be built and tested, the system responses are widely varying depending  on  the
                    particular design.

         Frequency response — Most noise exposure meters meet the allowable tolerances for a  Type
                    2 sound level meter (relative combined frequency response of microphone plus elec-
                    tronics) as specified in American National Standard Specifications for  Sound Level
                    Meters, SI.4-1971.

         Crest factor capability — Most devices can handle only small crest factors. Whether or not
                    this presents a problem depends on the use situation.  The response for all models,
                    with the exception of model D, falls below a 90% reading or exceeds a  110%
                    reading at a crest factor of 2-4.  Model D's response remains  nearly  perfect
                    at a crest factor of 4.

      •  Exchange rate — The exchange rate circuitry appears to be a troublesome  design  problem for
                    some manufacturers.  One reason for this may be that they have no experience with
                    such circuitry from other instruments; however, it is a crucial part  of a  noise
                    exposure meter.

      •  Temperature range — Most noise exposure meters suffer only a few percent error  due  to temp-
                    erature effects over the range 5°C-45°C.

         Battery voltage — Those devices with voltage regulation showed no effect in dosimeter reading
                    due to battery drain effect for voltages above the battery check indicator minimum.

      The obvious exception to the above general conclusions are models A and G, both of  which performed
poorly in each of the above tests.

      It is quite evident that a comprehensive performance standard for these devices is  an absolute
necessity.  American National Standards Institute Working Group S1-W45 is presently working on such
a standard and its efforts should be encouraged and accelerated.  In addition, a usage standard might  be
necessary to provide guidance on such items as microphone placement on the body, minimum recommended
checks prior to usage, and guideline handling procedures — Important considerations which the per-
formance standard may not provide.

      The test program has shown that there exists a wide variation in performance among the various
noise exposure meters tested.  Some might serve as instruments for monitoring compliance with the
occupational noise exposure regulation; however,  the user should be cautioned to carry out enough
evaluation tests  to ascertain that the devices are performing adequately for his purpose.
                                                   28

-------
                                           6.  BIBLIOGRAPHY


1.  American Standard Method for the Calibration of Microphones,  SI.10-1966,American National Standards
    Institute, New York, New York (March 1966).

2.  American National Standard Specifications for Sound Level Meters, SI.4-1971, American National
    Standards Institute, New York,  New York (April 1971).

3.  International Electrotechnical  Commission Recommendation on Precision Sound Level Meters, Publication
    179-1965, International Electrotechnical Commission, Geneva,  Switzerland (1965).

4.  American National Draft Standard for Integrating Sound-Level Meters, American National Standards
    Institute, New York, New York.

5.  Magrab, Edward B. and Blomquist, Donald S.,  The Measurement of Time-Varying Phenomena-Fundamentals
    and Applications. Wiley Intersclence, New York, New York (1971).
                                                  29

-------
NBS.114A (REV 7-731
    U.S DEPT. OF COMM.
   BIBLIOGRAPHIC DATA
         SHEET
1. PUBLICATION OR RKPORT NO
 NBSIR  73-417
 EPA-550/9-73-007
2. Gov't Accession
  No.
3. Recipient's Accession No.
4. TITLE AND SUBTITLE
   Evaluation of  Commercial Integrating-Type Noise Exposure
   Meters
                                               5. Publication Dace
                                                                        6. Performing Organization Code
7.
         am A. Leasure, Jr.,  Ronald  L.  Fisher, Marilyn A. Cadoff
                                                                                     i. Report No.
9. PERFORMING ORGANIZATION NAME AND ADDRESS

              NATIONAL BUREAU OF STANDARDS
              DEPARTMENT OF COMMERCE
              WASHINGTON, D.C. 20234
                                               10. Pro|eci/Task/W0rk Unu No.
                                                   2130491
                                               11. Contract/Grant No.
12. Sponsoring Organization Name and Complete Address (Street, City, State, ZIP)
   Office Noise Abatement  and Control
   U.  S.  Environmental Protection Agency
   Washington, D.  C.  20460
                                               13. Type of Report & Period
                                                 Covered

                                                     Final
                                               14. Sponsoring Agency Code
15. SUPPLEMENTARY NOTES
16. ABSTRACT (A 200-word or less factual summary ol most significant information  If document includes a significant
   bibliography or literature survey, mention it here )
  As  a result of the promulgation  of  occupational noise exposure regulations by the
  Federal government, there are a  number of  commercial noise exposure  meters on the
  market today  that provide a measure of noise integrated (with appropriate weighting)
  over a time interval.   This report  presents  the results of an evaluation of such
  instruments by the National Bureau  of Standards (under  the sponsorship of the U.S.
  Environmental Protection Agency) as to their usefulness in monitoring compliance
  with occupational noise regulations as well  as their applicability as instruments
  for use in achieving  the broader goals of  the EPA.  Tests  were designed and conducted
  to  evaluate microphone and system response to sound of  random incidence, frequency
  response,  crest factor capability,  accuracy  of the exchange rate  circuitry, perform-
  ance of the noise exposure meter as a function of temperature, and the dependence  of
  the device on battery voltage.   The rationale of the test  procedures utilized to
  evaluate overall system as well  as  specific  performance attributes,  details of  the
  measurement techniques, and results obtained are discussed.
17. KEY WORDS (six to twelve entries, alphabetical order, capitalize only the first letter of the first key word unless a proper
   name, separated by semicolons)
  Acoustics  (sound); dosimeter;  environmental acoustics;  instrumentation; noise
   exposure;  noise exposure meters.
18. AVAILABILITY           fX1 Unlimited


   I  ! For Official Distribution.  Do Not Release to NTIS


   I	I Order From Sup. of Doc., U.S. Government Printing Office
      Washington, D.C. 20402, SD Cat. No. C13	
   I  | Order From National Technical Information Service (NTIS)
      Springfield, Virginia 22151
                                    19. SECURITY CLASS
                                       (THIS REPORT)
                                       UNCLASSIFIED
                                     20. SECURITY CLASS
                                       (THIS PAGE)

                                       UNCLASSIFIED
                           21. NO. OF PAGES

                                 34-
                           22. Price
                                                                                  USCOMM.DC 20042-P74

-------