EPA-600/4-86-024
Method 200.6 - Dissolved Calcium, Magnesium, Potassium, and Sodium in
Net Deposition by Flame Atomic Absorption Spectrophotometry

[METHOD ONLY]
    EPQ
    1*00
    M
    ft*
      CO
                                               US EPA
                                    Headquarters and Chemical Libraries
                                        EPA West Bldg Room 3340
                                            Mailcode 3404T
                                        1301 Constitution Ave NW
                                          Washington DC 20004
                                             202-566-0556

-------
                                                     US EPA
                                       Headquarters and Chemical Libraries
                                            EPA West Bldg Room 3340
                                                 Mailcode 3404T
                                            1301  Constitution Ave NW
                                              Washington DC 20004
                                                  202-566-0556
                       Method 200.6 — Dissolved Calcium, Magnesium,  Potassium,
                                      and Sodium in Wet Deposition by Flame Atomic
                                      Absorption Spectrophotometry
                                          March 1986
   CO
   LLJ
Q_

                                  Performing Laboratory:

                                   Loretta M. Skowron
                                    Carla Jo Brennan
                                     Mark E. Peden

                               Illinois State Water Survey
                                Analytical Chemistry Unit
                                   2204 Griffith Drive
                                Champaign, Illinois 61820
                                                  U.S. Environmental Protection Agency
                                                  Region 5, UbMfyWiZJ)
                                                  H w«st Jackson Boulevard, 12th Roar
                                                  Chicago, IL 60604-3590
                                       Sponsoring Agency:


                                 John D. Pfaff,  Project Officer

e~}
^~"^                                Inorganic Analysis Section
                               Physical and Chemical Methods Branch
                          United States Environmental Protection Agency
                               Office of Research and Development
                         Environmental Monitoring and Support Laboratory
                                     Cincinnati, Ohio 45268
                                           200.6-1

-------
                                    INDEX
Section
Number                             Subject

   1                           Scope and Application
   2                           Summary of Method
   3                           Definitions
   4                           Interferences
   5                           Safety
   6                           Apparatus and Equipment
   7                           Reagents and Consumable Materials
   8                           Sample Collection, Preservation, and Storage
   9                           Calibration and Standardization
  10                           Quality Control
  11                           Procedure
  12                           Calculations
  13                           Precision and Bias
  14                           References
                                    TABLES

    Method Detection  Limits  and Concentration Ranges  for Flame Atomic
    Absorption  Spectrophotometric Analysis of Wet Deposition.
    Operating Conditions  and Suggested Calibration  Standard Concentrations
    for  the  Determination of Calcium, Magnesium, Potassium, and Sodium  in Wet
    Deposition  Samples.
    Single-Operator Precision and Bias for Calcium, Magnesium, Potassium, and
    Sodium Determined from Analyte  Spikes of Wet Deposition Samples.
    Single-Operator Precision and Bias for Calcium, Magnesium, Potassium, and
    Sodium Determined from Quality  Control Check Samples.
                                    FIGURES

 1.   Percentile Concentration Values Obtained from Wet  Deposition  Samples:
     Calcium,  Magnesium,  Potassium,  and Sodium.
                                   200.6-2

-------
1.   SCOPE AND APPLICATION

    1.1  This method is appiicaoie to the determination of calcium,
         magnesium, potassium, and sodium in wet deposition by flame atomic
         absorption spectrophotometry (FAAS).

    1.2  The term "wet deposition" is used in this method to designate rain,
         snow, dew, sleet, and hail.

    1.3  The method detection limits (MDL) for the above analytes determined
         from replicate analyses of quality control check solutions containing
         0.053 mg/L calcium, 0.018 mg/L magnesium, 0.012 mg/L sodium, and
         0.013 mg/L potassium are 0.007, 0.002, 0.003, and 0.003 mg/L,
         respectively.  The concentration range of this method is outlined  in
         Table 1.

    1.4  Figure  1 represents cumulative frequency percentile concentration
         plots of calcium, magnesium, potassium, and  sodium obtained from the
         analysis of over five thousand wet deposition samples.  These data
         should  be considered during the selection of appropriate calibration
         standard concentrations.

2.  SUMMARY OF METHOD

    2.1  A solution containing the element(s) of  interest  is aspirated as  a
         fine mist into a flame where it  is converted to an atomic  vapor
         consisting of ground state  atoms.  These ground state atoms  are
         capable of absorbing electromagnetic radiation over a series of
         very narrow, sharply defined wavelengths.  A distinct line source  of
         light,  usually a hollow  cathode  lamp specific to  the metal of
         interest, is used  to pass  a beam  through the flame.  Light from the
         source  beam,  less  whatever  intensity was absorbed by the  atoms  of  the
         metal of  interest,  is isolated by  the  monochromator and measured  by
         the  photodetector.   The  amount of  light  absorbed  by  the analyte is
         quantified by comparing  the light  transmitted through the flame to
          light transmitted  by a  reference  beam.   The  amount of  light absorbed
          in  the  flame  is  proportional  to  the  concentration of  the  metal  in
         solution.  The  relationship between  absorption  and concentration  is
         expressed by  Beer's Law:

                             log(I /I)  = abc = A
                                  o

         where:   I  = incident  radiant power
                  1°  = transmitted radiant power
                  a  = absorptivity (constant for a given system)
                  b  = sample path length
                  c  = concentration of absorbing species (mg/L)
                  A  = absorbance

          The atomic  absorption  spectrophotometer is  calibrated with standard
          solutions containing known concentrations of the element(s) of
          interest.  Calibration curves are constructed from which  the
          concentration of each analyte in the unknown sample is determined.
                                   200.6-3

-------
3.  DEFINITIONS

    3.1  ABSORBANCE (A) — the logarithm to the base ten of the reciprocal
         of the transmittance, (T):

                             A = log(l/T)

                      0.0044 A = the absorption of 1% of
                                 the transmitted light.

         The absorbance is related to the analyte concentration by Beer's Law
         (Sect. 2.1) where 1/T =1/1
                                  o

    3.2  ATOMIC ABSORPTION — the absorption of electromagnetic radiation by
         an atom resulting in the elevation of electrons from their ground
         states to excited states.  Atomic absorption spectrophotometry
         involves the measurement of light absorbed by atoms of interest as a
         function of the concentration of those atoms in a solution.

    3.3  SPECTRAL BANDWIDTH ~ the wavelength or frequency interval of
         radiation leaving the exit slit of a monochromator between limits set
         at a radiant power level half way between the continuous background
         and the peak of an emission line or an absorption band of negligible
         intrinsic width  (14.1).

    3.4  SPECTROPHOTOMETER — an instrument that provides the ratio, or a
         function of the ratio, of the radiant power of two light beams as a
         function of spectral wavelength.  These two beams may be separated
         in time and/or space.

    3.5  For definitions of other terms used in this method, refer to the
         glossary.  For an explanation of the metric system including units,
         symbols, and conversion factors see American Society for Testing and
         Materials  (ASTM) Standard E 380, "Metric Practices" (14.2).

4.  INTERFERENCES

    4.1  Chemical interference is the most frequently encountered
         interference  in atomic absorption spectrophotometry.  A chemical
         interference may prevent, enhance, or suppress the formation of
         ground state  atoms in the flame.  For example, in the case of
         calcium determinations, the presence of phosphate or sulfate can
         result in the formation of a salt that hinders proper atomization of
         the solution when it is aspirated into the flame.  This decreases the
         number of free, ground state atoms in the  flame, resulting in  lowered
         absorbance values.  Aluminum can cause a similar interference  when
         measuring magnesium.  The addition of appropriate complexing agents
         to the sample solution reduces or eliminates chemical interferences
         and may increase the sensitivity of the method.
                                   200.6-4

-------
   4.2  Alkali metals such as  sodium  and  potassium may  undergo  ionization  in
        an air-acetylene  flame  resulting  in  a  decrease  in  ground  state  atoms
        available  for measurement  by  atomic  absorption.  Addition of  a  large
        excess of  an easily  ionizable element  such as cesium will eliminate
        this  problem, since  cesium will be preferentially  ionized.  The
        preferential ionization of the cesium  solution  results  in an  enhanced
        atomic absorption signal for  both potassium  and sodium  (14.3).

   4.3  If a  sample containing low concentrations of the metal  being
        measured  is analyzed immediately  after a sample having  a
        concentration exceeding the highest  calibration standard, sample
        carry-over will  result in elevated  readings. To prevent  this
        interference, routinely aspirate  water (Sect.  7.2) for  about  15
        seconds  after a  high concentration  sample.   Depending on  the
        concentration of metal in the last  sample analyzed,  it  may be
        necessary to  rinse for longer time  periods.   Complete purging of the
        system is ascertained by aspirating  water until the absorbance
         readout  returns  to the baseline.

    4.4  Wet  deposition  samples are characterized by  low ionic strength and
         rarely contain  enough salts to cause interferences due to
         nonspecific  background absorbance.   The use  of background correction
         techniques is  not necessary and will decrease the  signal to noise
         ratio and lessen precision.

5.  SAFETY

    5.1   The  calibration standards, sample types, and most reagents used in
         this method pose no hazard to the analyst.  Use a fume hood,
         protective clothing, and  safety glasses when handling concentrated
         hydrochloric acid (Sect.  7.5-6).

    5.2  Use a fume hood, protective  clothing, and safety  glasses when
         preparing the lanthanum solution.   The  reaction between  the  lanthanum
         oxide and acid  (Sect.  7.7) is extremely exothermic.

    5.3  A permanent ventilation system is required  to  eliminate  the  large
         quantity  of hot  exhaust gases produced  during  instrument operation.
         Since acetylene  is  a  flammable gas, take precautions when  using it.
         To avoid  explosions,  never pass  acetylene through copper or
         high-copper alloy  (brass,  bronze) fittings  or  piping.

    5.4  The  operator must wear safety glasses to avoid eye  damage  from the
         ultraviolet light emitted by the flame.

    5.5  To avoid  in-line explosions, do  not allow the  pressure of acetylene
         being delivered  to  the instrument to  exceed 15 psig (10.6 g/m ).
         In the event of  a  flashback, turn off the gas  control  switch,  the
         instrument power, and the gas tanks.

    5.6  Follow manufacturer's operating  guidelines  carefully when optimizing
         gas  flow rates.   Too low gas flow  rates can result in  a combustion
         within  the gas  mixing chamber and  therefore a  flashback.
                                   200.6-5

-------
    5.7  Check that the drain tube from the gas mixing chamber, fitted with a
         safety trap,  is filled with water before igniting the flame.  Keep
         the drain tube filled to prevent explosion in the chamber.  The
         safety trap may be either looped or valved.

    5.8  Avoid any contact with a hot burner head.  Serious tissue burns will
         result.

    5.9  Follow American Chemical Society guidelines regarding safe handling
         of chemicals used in this method  (14.4).

6.  APPARATUS AND EQUIPMENT

    6.1  ATOMIC ABSORPTION SPECTROPHOTOMETER — Select a double-beam
         instrument having a dual grating monochromator, p'hotodetector,
         pressure-reducing valves, adjustable spectral bandwith, wavelength
         range of 190-800 nm, and provisions for  interfacing with a strip
         chart recorder or a suitable data system.

         6.1.1  Burner  ~ Use a  long path, single slot air-acetylene burner
                head supplied by the manufacturer of  the spectrophotometer.

         6.1.2  Hollow  Cathode Lamps ~ Single element lamps are  recommended.
                Multi-element  lamps are available but are  not  recommended.
                They generally have a shorter  lifespan, are less  sensitive,
                require a higher operating current, and increase  the chances
                of  spectral  interferences.  When  available, electrodeless
                discharge  lamps  (EDL) may  also be used.

         6.1.3  Monochromator  — To  increase sensitivity  of calcium and
                potassium determinations,  use  a  monochromator  equipped with a
                blaze  grating  in the range of  500-600 nm  (14.5).   For  the
                analysis of  sodium  and  magnesium, a blaze  grating in the range
                of  200-250  nm  is  adequate.

         6.1.4  Photomultiplier  Tube — A wide spectral range  (160-900 nm)
                phototube  is recommended.  Select a red-sensitive phototube  to
                detect potassium at 766.5 nm and to increase  sensitivity to
                calcium at  422.7 nm.

     6.2  The  first time any glassware  is used  for making  stock solutions and
         standards, clean  with 0.6  N HCl and  rinse  thoroughly with water
          (Sect.  7.2)  before use. Maintain a  set of Class A volumetric flasks
         to be used only when making dilute working standards for the  analysis
         of wet deposition samples.  Store filled with  water  (Sect. 7.2) and
         covered.
                                   200.6-6

-------
    6.3   LABORATORY FACILITIES — Laboratories used for the analysis of
         wet deposition samples should be free from external sources of
         contamination.  The use of laminar flow clean air workstations is
         recommended for sample processing and preparation to avoid the
         introduction of airborne contaminants.  If a clean air bench is
         unavailable, samples must be capped or covered prior to analysis.  A
         positive pressure environment within the laboratory is also
         recommended to minimize the introduction of external sources of
         contaminant gases and participates.  Windows within the laboratory
         should be kept closed at all times and sealed if air leaks are
         apparent.  The use of disposable tacky floor mats at the entrance to
         the laboratory is helpful in reducing the particulate loading within
         the room.

7.   REAGENTS AND CONSUMABLE MATERIALS

    7.1   PURITY OF REAGENTS — Use chemicals of reagent grade or better for
         all solutions.  All reagents shall conform to the specifications of
         the Committee on Analytical Reagents of the American Chemical Society
         (ACS) where such specifications are available.

    7.2   PURITY OF WATER ~ Use water conforming to ASTM Specification D
         1193, Type  II  (14.6).  Point of use 0.2 micrometer filters are
         recommended for all faucets supplying water to prevent the
         introduction of bacteria and/or ion exchange resins into reagents,
         standard solutions, and  internally formulated quality control check
         solutions.

    7.3   ACETYLENE  (C H ) — Fuel — Minimum acceptable acetylene purity
         is 99.5%  (v/v).  Change  the cylinder when the pressure reaches
         75 psig  (53 g/m }  if  the acetylene is packed in acetone.
         Pre-purified grades that contain a proprietary solvent can be used to
         30 psig  (21 g/m )  before replacement.  Avoid introducing these
         solvents  into  the  instrument.  Damage to  the instrument's  plumbing
         system can  result.  Solvent in the system is indicated by  abnormally
         high pulsating background noise.  To  prevent solvent  carryover,  allow
         acetylene cylinders to  stand for at  least 24 hours before  use.

         CAUTION:  Acetylene  is  a highly flammable gas.  Follow the
         precautions in Sect.  5.3-6  regarding  safe operating pressures,
         suitable  plumbing, and  operator safety.

    7.4  CESIUM SOLUTION  (1.0  mL =  100.0 mg Cs)  —  lonization  Supgressant —
         Dissolve  126.7 g of  cesium  chloride  (CsCl),  dried at  105 C for  one
         hour, in  water  (Sect.  7.2)  and  dilute to  1  L.  Store  at  room
         temperature in a high density polyethylene  or  polypropylene
         container.  Add  to samples  and  standards  as directed  in  Sect.  9.4 and
         11.4  for  the  determination  of potassium and sodium.

    7.5  HYDROCHLORIC  ACID  (6.0 N)  — Carefully  add  1  volume  of  concentrated
         hydrochloric  acid  (HC1,  sp  gr 1.19)  to  an equal  volume  of  water
          (Sect.  7.2).
                                   200.6-7

-------
7.6  HYDROCHLORIC ACID {0.6 N) —• Add 50 mL of concentrated hydrochloric
     acici (HC1, sp gr 1.195 to 900 mL of water (Sect. "7.21 and dilute to
     1 L.

7.7  LANTHANUM SOLUTION (1.0 mL = 100.0 mg La) — Releasing Agent — In a
     glass 1 L volumetric  flask, place 117.0 g of lanthanum oxide
     (La O ), dried at 105°C for one hour.  Add 6 N HCl very
     carefully to the solid in increments of about 0.5 mL.  Cool the
     solution between additions.  Continue adding the acid solution to the
     flask in increasing increments until a total of 500 mL of 6 N HCl has
     been added.  Dilute to 1 L with water  (Sect. 7.2).  Store at room
     temperature in a high density polyethylene or polypropylene
     container.  Add to samples and standards as directed in Sect. 9.4.3
     and 11.4 for the determination of calcium and magnesium.

     CAUTION:  Dissolving  lanthanum oxide in hydrochloric acid is a
     violently exothermic  reaction; use extreme caution when dissolving
     the reagent.  Refer to Sect. 5.2 for proper safety precautions when
     preparing this solution.

7.8  OXIDANT  (air) — The  air may be provided by a compressor or
     commercially bottled  gas supply.  Remove oil, water, and other
     foreign matter from the  air using a  filter recommended by the
     manufacturer.  Refer  to  the manufacturer's guidelines  for recommended
     delivery pressure.

7.9  STOCK STANDARD SOLUTIONS — Stock standard solutions may be
     purchased as certified solutions or  prepared from ACS  reagent  grade
     materials as detailed below.  Store  the  solutions at room temperature
      in  high density polyethylene or polypropylene containers.

       7.9.1  Calcium Solution,  Stock  (1.0 mL  =  1.0 mg Ca) — Add  2.497 g
             of  calcium carbonate  (CaCO  ),  dried at  180 C for one
             hour, to approximately 600  mL  of water  (Sect.  7.2).   Add
             concentrated  hydrochloric acid 
-------
8.  SAMPLE COLLECTION, PRESERVATION, AND STORAGE

    8.1  Collect samples in high density polyethylene (HOPE) containers  that
         have been thoroughly riiibed with ASTM Type II water (7.2).  Do  not
         use strong mineral acids or alkaline detergent solutions for cleaning
         collection vessels.  Residual acids may remain in the polyethylene
         matrix and slowly leach back into the sample.  Alkaline detergents
         may also leave residues that may affect the sample chemistry.   Cap
         collection bottles after cleaning to prevent contamination  from
         airborne contaminants; air dry collection buckets in a laminar  flow
         clean air workstation and wrap in polyethylene bags prior to use.  If
         a laminar flow workstation is not available, pour out any residual
         rinse water  and bag the buckets immediately.  Do not dry the bucket
         interior by  any method other than air drying in a laminar flow  clean
         air workstation.

    8.2  The frequency of sample collection and the choice of sampler design
         are dependent on the monitoring objectives.  In general, the use  of
         wet-only samplers  is recommended to exclude dry deposition
         contributions, minimize sample contamination, retard evaporation,
         and enhance  sample stability.  Sample collection frequency  may  vary
         from subevent to monthly sampling periods.  Collection periods  of
         more than one week are not recommended since sample integrity may be
         compromised  by longer exposure periods.

    8.3  The dissolution of particulate materials  can affect the  stability of
         calcium, magnesium, sodium,  and potassium in wet deposition samples
          (14.7).  This instability  generally  results  in  a concentration
          increase for these constituents.  Measurements  should  be made
         immediately  after  sample collection  to obtain representative  data.
         Refrigeration of samples at  4°C will  minimize but  not  eliminate
         concentration changes.

         8.1.1   Filtration  of  samples through  a 0.45  micrometer membrane leached
                 with  water  (Sect.  7.2)  is effective at stabilizing samples that
                 are  influenced by  the dissolution  of  alkaline  particulate matter
                 (14.7).  Monitoring of  the  filtration procedure is necessary to
                 ensure  that samples are  not  contaminated by the membrane or
                 filtration  apparatus.  Filtered  samples  are stable for six weeks
                 when  stored at  room temperature.

 9.  CALIBRATION AND  STANDARDIZATION

    9.1   SETTING INSTRUMENT PARAMETERS

          9.1.1   Lamp Current  —  Refer to manufacturer's  guidelines for
                 optimization  of  this  parameter.   The use of excessively high
                 currents  will  shorten lamp life.   High  currents also cause
                 line broadening,  resulting in a reduction in sensitivity and
                 calibration curve linearity,  especially  in the determination of
                 magnesium.   The use of currents that are too low will cause lamp
                 instability and insufficient throughput  of energy through the
                 instrument's optical system.  The result is increased signal
                 noise due to excess electrical gain applied to the photodetector.


                                   200.6-9

-------
9.1.2  Light Beam — Position a small card over the burner slot to
       intercept the light beam from the hollow cathode lamp.  Check
       that the beam is focused midway along the slot and, if
       necessary, focus according to the manufacturer's guidelines.
       Rotate the lamp within its holder for maximum energy output
       readings.

9.1.3 Burner/Beam  Alignment — Position & small card  over the burner
       slot to  intercept the  light  beam from  the hollow cathode lamp.
       For optimal sensitivity when analyzing calcium,  magnesium,
       potassium, and sodium, adjust the burner height  so that the
       center of the light beam is approximately 6  mm above the
       surface  of  the  burner slot.  By adjusting the burner alignment
       and rotation, set the light beam to coincide with the burner
       slot.  While observing from above, move the  card along the
       full length of the burner slot to ensure that the beam is
       centered over the slot for the entire length of the burner.
       Optimize this parameter for maximum instrumental sensitivity
       as directed in Sect. 9.2.

9.1.4  Wavelength — Set the wavelength of the spectrophotometer for
       each analyte according to Table 2 by following the
       manufacturer's  operating guidelines. After  the  instrument has
       warmed up with  the  flame  burning (about 30  minutes), check the
       wavelength and readjust if necessary.

       Note:  The sodium spectrum is characterized by a doublet at
       589.0 nm and 589.5 nm.  The wavelength chosen for sodium
       determinations  depends  on  the degree of analytical sensitivity
       desired by the operator.  A setting of 589.0 nm will provide
       maximum sensitivity in the concentration range of most wet
       deposition samples.  For those samples with higher sodium
       concentrations, a less sensitive setting of 589.5 nm is more
       appropriate.  Refer to Tables 1 and 2 for information
       regarding working ranges, standards, and detection limits for
       sodium at each wavelength setting.

9.1.5  Spectral Bandwidth — The selection of optimum bandwidth
       depends  upon the spectrum of  the particular element being
       analyzed.  For the determination of calcium, magnesium, and
       potassium, a relatively wide  (1.0 nm) bandwidth  is
       appropriate.  Because the sodium spectrum is characterized by
       a doublet, use a smaller bandwidth of 0.5 nm.

9.1.6  External Gas Settings — Follow manufacturer's recommended
       delivery pressures for air and acetylene.  Never allow
       acetylene pressure to exceed  15 psig (10.6 g/m  ).
                          200.6-10

-------
    9.1.7  Nebulization  Rate  —  Set the acetylene  and  air  flow  rates  as
           recommended by  the manufacturer.  Adjust  the  nebulizer  sample
           uptake rate to approximately  5 mL/min. If an adjustable  glass
           bead nebulizer  is  used, adjust it according to  manufacturer's
           guidelines.   Exact placement of the  glass bead  is  critical to
           ensure that a uniform vapor of the smallest size particles is
           introduced into the flame.  Improper spacing  of the  bead from
           the  nebulizer end  will result  in poor precision and
           sensitivity.  Optimize the sample uptake  rate for  maximum
           sensitivity as  directed in Sect. 9.2.

           Note:  The nebulizer  can clog  easily if particulates are
           present in the  samples.  Symptoms of this are decreased
           sensitivity and/or dramatically increased signal noise,
           especially noticeable at the higher  concentration  levels.
           A thorough cleaning with a small diameter wire is  usually
           sufficient to unclog the nebulizer.

     9.1.8 Flame Conditions — If the flame temperature  is too  low,
           compounds containing the analyte will not be  completely
           dissociated.   Alternatively, too high a flame temperature  may
           result in ionization.  In both cases, a decrease  in  the
           apparent concentration of the  analyte will result.  In
           general, calcium exhibits maximum  sensitivity at  higher fuel
           and oxidant flow rates. Maximum sensitivity  for  potassium is
           obtained with minimal gas flow rates,  resulting  in  lower  flame
           temperature and allowing  longer residence time of  the atomic
           vapor in the  flame.  The MOLs  stated in Sect. 1.3  for
           magnesium and sodium are obtained  over a  wide range  of flame
           conditions.   Optimize this parameter for  maximum instrumental
            sensitivity as directed in Sect.  9.2.

           CAUTION:  Follow manufacturer's operating guidelines
            carefully when setting gas  flow rates since combustion within
            the gas mixing chamber can occur  if  caution is not exercised.

9.2 Optimization — Allow  the instrument  to warm  up for 30 minutes before
     beginning  the  optimization. Set  the  instrument readout  to absorbance
     units and set the integration time  to <0.5  seconds.   Use either a
     strip chart recorder or set the display  in  a continuous read mode to
     monitor  absorbance readings.  Aspirate a  calibration standard at a
     concentration near the midpoint of  the working  range  (Sect. 9.43.
     While watching the absorbance readings,  adjust  the  instrument
     parameters with small, discrete changes  until maximum values are
     obtained.  Parameters such as flame conditions,  nebulization rate,
     and the region of maximum atom concentration  in the flame  are
     interrelated.  Adjustment of any of these  three parameters usually
     requires adjustment of the other two.
                              200.6-11

-------
9.3  Instrument Response Time — Determine the minimum sample uptake time
     before taking  a  reading on a sample or standard  solution.  Use either
     ft strip chart recorder or set the display in a continuous read mode
     to monitor absorbance readings.  After purging the system with water
     (Sect. 7.2), aspirate the highest calibration standard (Sect. 9.4)
     and measure the length of time necessary to obtain a stable reading.
     Aspirate water fSect. 7.2) and measure the time it takes for the
     baseline to return to zero.

     Note:  If the time necessary for the baseline to return to zero is
     longer than 15 seconds, a clogged nebulizer may be suspect.  If
     purging  tine begins  co  increase  during sample analysis,  this may also
     be an indication of nebulizer clogging.

9.4  CALIBRATION SOLUTIONS

     9.4.1 Five  calibration solutions and  one  zero  standard are needed to
            generate  a  suitable calibration curve.  The  lowest  calibration
            solution should  contain the analyte of interest at a
            concentration greater than or equal to the method detection
            limit.  The highest solution should approach the expected
            upper  limit of  concentration of the analyte in wet deposition.
            Prepare the remaining solutions such that  they are evenly
            distributed throughout the concentration range.  Suggested
            calibration standards for each  analyte are listed in Table  2.

     9.4.2  Prepare all calibration standards by diluting the stock
            standards  (Sect.  7.9) with water  (Sect.  7.2).  Use glass
             (Class  A) or plastic pipettes that  are within the bias and
            precision tolerances specified  by the manufacturer.  The
            calibration  standards are stdble for three months if stored at
            room temperature in high  density polyethylene or polypropylene
            containers.

     9.4.3  After  preparing  the calibration standards  to volume,  add  the
             lanthanum solution (Sect. 7.7)  to the  calcium and magnesium
             standards to  yield 1000 mg/L La.  Add  the  cesium  solution
             (Sect.  7.4)  to  the potassium and sodium  standards for
             1000 rag/L Cs.   Mix well.  Use the same stock of  xonization
             suppressant  or  releasing  agent  for  the samples  and  the
             calibration  standards.

             Note:   The  final volume of  each working  standard  solution
             exceeds the  nominal volume  toy 1%.   This  adjustment  is
             necessary  to  maintain  consistency when the appropriate  volume
             of suppressor solution is added to the  wet deposition  samples.
                               200.6-12

-------
9.5  CALIBRATION

         9.5.1  To establish a baseline, aspirate the zero standard and set
                the absorbance readout to 0.000.  Aspirate the calibration
                standards, allowing tints for each standard to equilibrate in
                the flame and gas mixing chamber before measuring the
                absorbance  (Sect. 9.3).  Construct calibration curves for each
                of the four analytes according to Sect. 12.

         9.5.2  Analyze all the calibration standard solutions.  The apparent
                concentration values must agree with the nominal
                concentrations within the predetermined control limits  (Sect.
                10.2.1) of three times the standard deviation  (±3s) .  If
                results fall outside of these limits, recalibrate the
                instrument.  If there is a consistent bias greater  than
                x _* 2s and less than x +_ 3s, for all of the concentration
                values measured, reestablish the baseline with the  zero
                standard and reanalyze the calibration standards.

         9.5.3  Verify the calibration curve after every ten samples and  at
                the end cf each day's analyses according to Sect. 10.7.

10.  QUALITY CONTROL

     10.1   Each laboratory using this method should develop formalized  quality
            control protocols to continually monitor the bias and precision  of
            all measurements.  These protocols are required to  ensure that the
            measurement system is in a state of  statistical control.  Estimates
            of bias and precision for wet deposition analyses cannot be
            made unless these control procedures  are followed.  Detailed
            guidelines  for  the development of quality assurance and  quality
            control protocols for precipitation measurement systems  are
            published in a  manual available  from  the United States
            Environmental  Protection Agency, Research Triangle  Park, NC   27711
            (14.8).  Included in  this manual are  procedures for the  development
            of statistical  control  charts  for use in monitoring bias and
            precision as well as  recommendations  for the  introduction of
            reagent blanks,  laboratory duplicates, field  duplicates, spike
            samples, and performance evaluation  samples.  These guidelines are
            to be  used  by  all laboratories  involved with  wet deposition
            measurements.

      10.2   ESTABLISHMENT  OF WARNING AND CONTROL LIMITS ~  Warning  and  control
            limits are  used to monitor  drift in  the  calibration curve,  analyses
            of quality  control check samples (QCS),  and measured  recoveries
            from  laboratory spikes.

            10.2.1 Calibration  Curve  — After  a calibration  curve  has been
                   constructed  according  to Sect.  12,  reanalyze  additional
                   aliquots of  all the  standards.   Calculate  the
                   concentrations  using the previously derived calibration
                   curve.   Repeat  this  procedure until at  least  ten
                                    200.6-13

-------
        determinations  at  each concentration level have been made.
        These  data should  be collected on ten different days to
        provide  a realistic estimate of the method variability.
        Calculate a standard deviation (s)  at each concentration
        level.   Use the nominal standard concentration as the mean
        value  (x) for determining the control limits.   A warning
        limit  of x ^ 2s and a control limit of x ^ 3s  should be
        used.   Reestablish these limits whenever instrumental
        operating conditions change.

10.2.2  Quality Control Check Samples {QCS} — Calculate warning
        and control limits for QCS solutions from a minimum of ten
        analyses performed on ten days.  Use the calculated
        standard deviation  (s) at each QCS concentration level to
        develop the limits as described in Sect. 10.2.1.  Use the
        certified or NBS traceable concentration as the mean
        (target) value.  Constant positive or negative measurements
        with respect to the true value are indicative of a method
        or procedural bias.  Utilize the data obtained from QCS
        measurements as in Sect. 10.6 to determine when the
        measurement system  is out of statistical control.  The
        standard deviations used to generate the QCS control  limits
        should be comparable to the single operator precision
        reported  in Table 4.  Reestablish new warning and control
        limits whenever instrumental operating conditions are
        varied or QCS concentrations are changed.

10.2.3  Laboratory Spike Solutions  — A minimum of ten analyte
        spikes of wet deposition samples is  required to develop a
        preliminary data base  for the calculation of warning  and
        control  limits for  spike recovery data.   Select the spike
        concentration such  that the working  range of  che method
        will not  be exceeded.  Samples  selected for the  initial
        spike recovery study  should represent the concentration
        range common to wet deposition  samples  in order  to  reliably
        estimate the method accuracy.   Calculate  a mean  and
        standard deviation  of  the percent  recovery data  using the
        formulas provided  in  the glossary.   Determine  warning and
        control  limits using  H-2s and +3s,  respectively.   If
        the data indicate  that no significant method  bias  exists
         (14.9),  the  100 percent  recovery  is used  as the mean
        percent  recovery.   Where a  significant  bias  is determined
        at the  95% confidence level,  the control  limits are
        centered around the bias estimate.   Routine  spiked sample
        analyses that  yield percent recovery data outside  of the
        control  limits are an indication of matrix interferences
        that  should be resolved  before routine  analyses are
        continued.
                         200.6-14

-------
      10.2.4  All warning and control limits should be reevaluated on a
              continual basis as additional data are collected during
              routine analyses.  The limits should be broadened c~
              narrowed if a recalculated standard deviation under similar
              operating conditions provides a different estimate of the
              procedure variability.

10.3  Monitor the cleaning procedure by pouring a volume of water (Sect.
      7.2}  that approximates the median sample size into the collection
      vessel.  Allow the water to remain in the sealed or capped
      collection container for at least 24 hours and determine the
      concentration of the analytes of interest.  If any of the measured
      concentrations exceed the MDL, a contamination problem is indicated
      in the cleaning procedure.  Take corrective action before the
      sampling containers are used for the collection of wet deposition.

10.4  Keep daily records of calibration data and the instrument operating
      parameters used at the time of data acquisition.  Use these
      historical data as general performance indicators.  Gross changes
      in sensitivity, curve linearity, or photomultiplier tube voltage
      are indicative of a problem.  Possibilities include instrument
      malfunction, clogged nebulizer, incomplete optimization, bad hollow
      cathode lamp, contamination, and inaccurate standard solutions.

10.5  Precision will vary over the analyte concentration range.  Standard
      deviation  (s) increases as concentration  increases while relative
      standard deviation  (RSD) decreases.  At approximately 100 times  the
      MDL,  the RSD should remain less than 1%.

10.6  Analyze a quality control check sample  (QCS) after a calibration
      curve has been established.  This sample may be formulated in  the
      laboratory or obtained from the National Bureau of Standards  (NBS
      Standard Reference Material 2694, Simulated Rainwater).  The check
      sample(s) selected must be within the range of the calibration
      standards.  Prepare according to Sect. 11.4.  If  the measured  value
      for the QCS falls outside of the +3s limits  (Sect. 10.2.2), or if
      two successive QCS checks are outside of the +2s  limits, a
      problem is  indicated with the spectrophotometer or calibration
      curve.  Reestablish the baseline with the  zero standard  and/or
      recalibrate.   If  the QCS  analysis is still beyond control  limits,
      inaccurate working standards might be the  problem.  Prepare new
      standards.  Plot  the data obtained from the QCS checks on  a control
      chart  for  routine assessments of bias and precision.

10.7  Verify the  calibration curve after a maximum  of ten  samples and  at
      the end of  each  day's analyses.  Analyze  a zero standard and
      calibration standards at  the  low and high ends of the working
      range.   If  the  routine calibration checks do  not  meet the  criteria
      described  in Sect.  10.6,  recalibrate the  system and  reanalyze  all
      samples  from the  last time the  system was in  control.  Verify  the
      new calibration  curve with  the  QCS according  to Sect. 10.6 and
      reanalyze  all  samples  from  the  last  time  the  measurement system  was
      in control.
                               200.6-15

-------
10.8  Submit a Field Blank (FB) to the laboratory for every 20 samples.
      The FB may consist of a water sample (Sect. 7.2) or a known
      reference solution that approximates the concentration levels
      characteristic of wet deposition.  The FB is poured into the
      sampling vessel at the field site and undergoes identical
      processing and analytical protocols as the wet deposition
      sample(s).  Use the analytical data obtained from the FB to
      determine any contamination introduced in the field and laboratory
      handling procedures.  The data from the known reference solution
      can be used to calculate a system precision and bias.

10.9  Prepare and analyze a laboratory spike of a wet deposition sample
      according to the guidelines provided in "Quality Assurance Manual
      for Precipitation Measurement Systems" (14.8).  Compare the
      results obtained from the spiked samples to those obtained from
      identical samples to which no spikes were added.  Use these data
      to monitor the method percent recovery as described in Sect.
      10.2.3.

10.10  Participation in performance evaluation studies is recommended for
       precipitation chemistry laboratories.  The samples used for these
       performance audits should contain the analytes of interest at
       concentrations within the normal working range of the method.  -The
       true values are unknown to the analyst.  Performance evaluation
       studies for precipitation chemistry laboratories are conducted
       semiannually by the USEPA Performance Evaluation Branch, Quality
       Assurance Division, Research Triangle Park, NC  27711.

10.11  INSTRUMENT MAINTENANCE — Strictly adhere to manufacturer's
       maintenance schedule.

       10.11.1  Exposed optical mirrors should be replaced yearly to
                maintain optimal sensitivity and precision.

       10.11.2  If the instrument is used for other sample types that
                have high analyte concentrations it nay be necessary to
                disassemble the entire burner-nebulizer system for
                cleaning before analyzing wet deposition samples.  This
                is best accomplished by placing the components in a water
                (Sect. 7.2) bath in an ultrasonic cleaner for a half
                hour.  Rinse with water (Sect. 7.2) after cleaning and
                allow to air dry in a dust-free environment before
                reassembly.  Check o-rings for wear and replace if
                necessary.
                              200.6-16

-------
11.  PROCEDURE

     11.1  Set instrument parameters and optimize the instrument each day
           according to Sect. 9.1-2.

     11.2  Prepare all standards and construct calibration curves according
           to Sect. 9.4-5.

     11.3  After the calibration curve is established, analyze the QCS.  If
           the measured value for the QCS is not within the specified limits
           (Sect. 10.2.2), refer to Sect. 10.7.

     11.4  Pipette the appropriate cesium or lanthanum solution into the
           empty sample cup  (Cs or La:Sample = 1:100).  For the determination
           of calcium and magnesium, use the lanthanum solution described in
           Sect. 7.7.  For potassium and sodium determinations, add cesium
           solution (Sect. 7.4).  Pour the sample into the sample cup
           containing Cs or La; 3 mL of sample for 30 uL of Cs or La is
           suggested.  Mix well, aspirate, wait for equilibration in the flame
           (Sect. 9.3), and record the measured absorbance (or concentration).

     11.5  If the absorbance (or concentration) for a given sample exceeds
           the working range of the system, dilute a separate sample with
           water (Sect. 7.2).  Prepare and analyze according to Sect. 11.4.

     11.6  When analysis is complete, rinse the system by aspirating water
           (Sect. 7.2) for ten minutes.  Follow the manufacturer's guidelines
           for instrument shut-down.

12.  CALCULATIONS

     12.1  For each analyte of interest, calculate a linear least squares fit
           of the standard concentration as a function of the measured
           absorbance.  The linear least squares equation is expressed as
           follows :
           where:  y  = standard concentration in mg/L
                   x  = absorbance measured
                   BQ = y-intercept calculated from:  ? - B 5c
                   B. = slope calculated from:

                    n                   n
                    £ (x  - x) (y  - y-)/ £  (x  - x}
                   i-l                 1=1

                         where:  x = mean of absorbances measured
                                 y = mean of standard concentrations
                                 n = number of samples

           The correlation coefficient should be 0.9995 or greater.  Determine
           the concentration of analyte of interest from the calibration
           curve.
                                   200.6-17

-------
     12.2  If the relationship between concentration and absorbance is
           nonlinear,  use a second degree polynomial least squares equation to
           derive a curve with a correlation X).9995.  The second degree
           polynomial  equation is expressed as follows:

                                    y = B2x  + Bj^x + BQ

           A computer  is necessary for the derivation of this function.
           Determine the concentration of analyte of interest from the
           calibration curve.

     12.3  An integration system or internal calibration software may also
           be used to  provide a direct readout of the concentration of the
           analyte of  interest.

     12.4  Report concentrations in mg/L as Ca+ , Mg+ , Na ,  and K+.
           Do not report data lower than the lowest calibration standard.

13.  PRECISION AND BIAS

     13.1  The mean percent recovery and mean bias of this method were
           determined  from the analysis of spiked wet deposition samples
           according to ASTM Standard Practice D4210, Annex A4 (14.9).  The-
           results are summarized in Table 3.  No statistically significant
           biases were found for any of the metal cations.

     13.2  Single-operator precision and bias were obtained from the analysis
           o£ quality  control check samples that approximated the levels
           common to wet deposition samples.  These results reflect the
           accuracy that can be expected when the method is used by a
           competent operator.  These data are presented in Table 4.

1^.  REFERENCES

     14.1  Annual Book of ASTM Standards, Part 42, "Standard Definitions of
           Terms and Symbols Relating to Molecular Spectroscopy," Standard E
           131-81, 1981, p. 66.

     14.2  Annual Book of RSTM Standards. Section 11, Vol. 11.01  (1),
           "Excerpts from Standard for Metric Practice," Standard E 380-79,
           1983, pp. 679-694.

     14.3  Van Loon, J. C., Analytical Atomic Absorption Spectroscopy,
           Selected Methods Academic Press, Inc., New York, N. Y., 1980,
           p. 42.

     14.4  "Safety in Academic Chemistry Laboratories," American Chemical
           Society Publication, Committee on Chemical Safety, 3rd Edition,
           1979.
                                   200.6-18

-------
14.5  Instrumentation Laboratory, Inc., Operator's Manual Model IL951,
      AA/AE Spectrophotometer, Instrumentation Laboratory, Inc.,
      Wilmington, Massachusetts, 1982, pp. 3-4.

14.6  Annual Book of ASTM Standards, Section 11, Vol. 11.01 (1),
      "Standard Specification for Reagent Water," Standard D 1193-77,
      1983, pp. 39-41.

14.7  Peden, M. E. and Skowron, L. M., 'Ionic Stability of Precipitation
      Samples," Atmos. Environ. 12, 1978, pp. 2343-2349.

14.8  Topol, L. E., Lev-On, M., Flanagan, J., Schwall, R. J., Jackson, A.
      E., Quality Assurance Manual for Precipitation Measurement
      Systems, 1985, U.S. Environmental Protection Agency,
      Environmental Monitoring Systems Laboratory, Research Triangle
      Park, NC 27711.

14.9  Annual Book of ASTM Standards, Section 11, Vol. 11.01  (1),
      "Practice  for Intralaboratory Quality Control Procedures  and a
      Discussion of Reporting Low-Level Data," Standard D4210 Annex  A4,
      1983, pp.  15-16.
                               200.6-19

-------
     Table 1.  Method Detection Limits and Concentration Ranges for
               Flame Atomic Absorption Spectrophotometnc Analysis
               of Wet Deposition.
Analyte
Method Detection
     Limit/
      mg/L
Concentration
    Range,
     mg/L
Calcium
Magnesium
Potassium
Sodium
0.007
0.002
0.003
0.003d
0.030 -
0.010 -
0.010 -
0.010 -
3.00
1.00
1.00
i.ooa
                       0.007
                                                0.020 - 2.00
a.   589.0  nm wavelength  setting
b.   589.5  nm wavelength  setting
                             200.6-20

-------
     Table 2.   Operating Conditions and Suggested  Calibration
                        Concentrations for the  Determination of
               Calcium,  Magnesium,  Potassium,  and  Sodium  in  Wet
               Deposition Samples.
Analyte
Wavelength
 Setting,
    nm
 Spectral
Bandwidth,
    nm
 Working
Standards,
   mg/L
Calcium
   422.7
    1.0
                                                        zero
                                                        0.03
                                                        0.75
                                                        1.50
                                                        2.25
                                                        3.00
Magnesium
   285.2
    1.0
                                       zero
                                       0.01
                                       0.25
                                       0.50
                                       0.75
                                       1.00
Potassium
   766.5
    1.0
                                                        zero
                                                        0.01
                                                        0.25
                                                        0.50
                                                        0.75
                                                        1.00
Sodium
   589.0
    0.5
                                                        zero
                                                        0.01
                                                        0.25
                                                        0.50
                                                        0.75
                                                        1.00
                    589.5
                    0.5
                                                        zero
                                                        0.02
                                                        0.50
                                                        1.00
                                                        1.50
                                                        2.00
a.  Based on the MDL and 95th percentile concentration of each analyte
    obtained from analyses of over five thousand wet deposition samples
    from the NADP/NTN precipitation network.
b.  Refer to Sect. 9.1.2 for details on wavelength selection
                             200.6-21

-------
         Table  3.  Single-Operator  Precision and Bias  for Calcium,
                   Magnesium,  Potassium, and Sodium Determined  from Analyte
                   Spikes  of Wet  Deposition Samples.
Analyte
Calcium
Magnesium
Potassium
Sodium
Amount
Added,
mg/L n
0
0
0
0
0
0
0
0
.087
.221
.018
.045
.021
.052
.099
.249
20
20
20
20
18
13
19
20
Mean
Percent
Recovery
101
98
97
96
145
108
107
100
.5
.3
.2
.6
.2
.1
.1
.2
Mean
Bias,
mg/L
0.
-0.
-0.
-0.
0.
0.
0.
0.
001
003
001
002
010
004
007
000
Standard
Deviation,
mg/L
0
0
0
0
0
0
0
0
.010
.011
.001
.002
.006
.002
.011
.008
Statistically
Significant
Bias?
No
No
No
No
No
No
No
No
a.  Number of replicates
b.  95% Confidence Level
c.  589.0 nm wavelength
                                   200.6-22

-------
          Table 4.   Single-Operator Precision and Bias for Calcium,
                    Magnesium,  Potassium,  and Sodium Determined
                    from Quality Control Check Samples.
Theoretical
Concentration,
Analyte mg/L
Calcium
Magnesium
Potassium
Sodium
0
0
0
0
0
0
0
0
.053
.406
.018
.084
.021
.098
.082
.465
Measured
Concentration,
mg/L n
0
0
0
0
0
0
0
0
.051
.413
.017
.083
.020
.095
.084
.479
145
145
145
145
127
122
123
122
Bias,
mq/L %
-0
0
-0
-0
-0
-0
0
0
.002
.007
.001
.001
.001
.003
.002
.014
-3.8
1.7
-5.6
-1.2
-4.8
-3.1
2.4
3.0
Precision,
s , RSO ,
mg/L %
0
0
0
0
0
0
0
0
.002
.003
.001
.001
.001
.001
.001
.003
3.9
0.7
5.9
1.2
5.0
1.0
1.2
0.6
The above data were obtained from records of measurements made under the
direction of the NADP quality assurance program.

a.  Number of replicates
b.  589.0 nm wavelength
                                   200.6-23

-------
NJ
§

T
                                   Figure !„  Percentile  Concentration  Values Obtained  from
                                               Wet Deposition Samples:   Calcium, Magnesium,
                                               Potassium,  and Sodium.
                                                   calcium
i.oo
           2.00
                       3.00
                                                potassium
                            o.io
                                   0.20
                                                                     20
                                                                     10
                                                                                                 magnesium
                                                           0.40
                                                                       0.60
                                                                                                     sodium
                                                 0.40    0.50

                                                      CONCENTRATION (mg/L)
                                                                           0.50
                                                                                     1.50
                                                                                              2.50
                                                                                                        3.50

-------
EPA - Online Library System Record
wysiwyg://42/http://cave epa.gov/c ..6M%3D2%26K%3D85856%26R%3DY%26U%3D 1
                            Libraries
                        J
                           Contact Us I Print Version
                         f^EPAHpme > Information Sources > Libraries > Online Library System>OLS Recordr
        OLS Record
                  rr~~
                  *=
        Wh III' lerns" II ISuMHit all I PrlnT
                                          Display Records as Bibliography
        RECORD NUMBER: 2
Main Title
Author
CORP Author
Year Published
Report Number
Holdings
Owner Libraries
Stock Number
Document Type
Control Number
Cataloging Source
NTIS Title Notes
PUB Date Free Form
Collation
Supplementary Notes
Subjects
Category Codes
NTIS Prices
Primary Description
Abstract
t
Development of Standard Methods for the Collection and Analysis of
Precipitation
Peden, M E. ; Bachman, S R ; Brennan, C J , Demir, B. , James, K O. ;
Illinois State Water Survey Div , Champaign .Environmental Monitoring and
Support Lab.-Cmcinnati, OH.
1986
EPA-R-810780; EPA/600/4-86/024;
LIBRARY CALL NUMBER
MTIC MOST EPA LIBRARIES HAVE FICHE COPY. CHECK WITH
N ' lb INDIVIDUAL LIBRARY ABOUT PAPER COPY.
- National Technical Information Service
PB86-201365
NT
619921977
NTIS/MT
Rept. for May 83-Apr 86.
May 86
278p |
Sponsored by Environmental Monitoring and Support Lab.-Cincinnati, OH.
Standards, Precipitation(Meteorology); Water pollution. Evaluation; Collecting
methods; Manuals; Preserving, Processing; Accuracy, Flame photometry,
Colonmetric analysis; Spectrophotometry, Chromatographic analysis;
Volumetric analysis; Laboratories, Comparison; Acid precipitation
1 3B; 4B; 7D. 68A; 68D; 55C; 99A j
PCA13/MFA01 |
600/06
To provide the scientific community with a set of standardized procedures for '
the collection and analysis of wet precipitation samples, an analytical methods
manual for use in acid deposition studies has been developed The manual
includes detailed methods documentation for the major inorganic constituents
of interest in wet deposition as well as guidelines for the collection,
preservation, and processing of samples. The analytical methodologies include
 Iof2
                                                     4/21/03 11:52 AM

-------
EPA - Online Library System Record
wysiwyg-//42/http://cave.epa.gov/c.. 6M%3D2%26K°/o3D85856%26R%3DY^o26U%3DI
                                 ii preservation, and processing of samples. The analytical methodologies include
                                   flame atomic absorption spectrophotometry, ion selective electrode, automated
                                   colonmetry, ion chromatography, and titrimetnc procedures.
                      SimnariJ
                                  Heist
BElp
    Display Records as Bibliography
                                              If there is an AREA footer, put it here
                                        EPA Home I Privacy and Security Notice I Contact Us

                                          Last updated on Tuesday, December 31st, 2069
                                   URL http //cave epa gov/cgi/nph-bwcgis/BASIS/ncat/lib/ncat/DDW
 2 of 2
                                                                                                            4/21/03 11:52 AM

-------