Determination of Po1on1um-2lO
In Drinking Water by Alpha Particle  Counting


                Method  912.0
              Richard J.  VeHen
                     and
               Betty J.  Jacobs
          Inorganic Analyses Section
     Physical and Chemical Methods Branch
Environmental Monitoring  and  Support Laboratory
     U.S. Environmental Protection Agency
               Cincinnati, Ohio

                 October  1983

-------
                       Determination of Po1on1um-2lO 1n
                  Drinking Water by Alpha Particle Counting
                                 Method 912.0


1.  Scope and Application
    l.l  This method is applicable to the determination of polon1um-210
         (Po-210) In drinking water samples.
    1.2  The method detection limit as defined by the National Interim
         Primary Drinking Water Regulations 1s 0.1 pC1/L based on an
         Instrumental alpha particle background of 0.1 count /minute and a
         100-minute counting period.
    1.3  This method may be applied to surface and ground water samples
         provided these samples are Immediately filtered upon collection.

2.  Summary of Method
    771Lead is added as a nonisotopic carrier and the polonium-210 is
         concentrated by coprecipitatlon on lead sulfide from an acetic acid
         solution using hydrogen sulfide gas.  The lead sulfide is collected
         by filtration and dissolved  in concentrated nitric acid.  Nitrates
         are removed by fuming with a small quantity of perchloric acid.
         The solution  is treated with diluted hydrochloric acid and the
         polonium-210  1s spontaneously electroplated onto a nickel disk.
         The nickel  disk 1s washed, dried,  and alpha.counted.

3.  Safet
     371Hydrogen  sulfide  gas  is  not  only  toxic  but  also highly  flammable.
          Its use should  be restricted to well  ventilated hood facilities  and
        --away-from all open flames.
     3.2   Perchloric acid is used  in a very limited  amount.. Adequate
          precautions should be taken  to insure that  the acid fumes  are
          purged through  a water trap  when  not  used  1n hoods specifically
          designed  for perchloric  acid usage.  ~

 4.   Apparatus
     471Concentric ring water bath.
     4.2   Plating  cell, 8 ounce plastic nursing bottle (See Fig.  13.1).
     4.3.  Stirring motor  with glass stirrer.
     4.4   Internal  proportional counter, -
     4.5   Membrane  filtering assembly, 300  ml.

 5.   Reagents
     5.1   Acetic acid, glacial, CAS Reg 64-19^7,  99.7% w/w.
     5.2   Acetone,  CAS Reg 67-64-1.
     5.3   Anmonium hydroxide, CAS Reg 1336-21-6 (NfyOH) 6N.  Dilute 400 ml
          of concentrated  ammonium hydroxide to 1000 ml uling distilled water.
     5.4   Hydrochloric acid, CAS Reg 7647-01-0.  (HC1) 0.5N.   Dilute 42 ml
          36-381 HC1 to 1000 mL using distilled water.
     5.5  Hydrogen sulfide'gas  (H?S).   Lecture bottle 99.5%.
     5.6  Lead nitrate carrier  solution, CAS Reg 10099-74-8,  (Pb(N03>9)
          0.1N.  Dissolve  16.5g lead nitrate (Pbfffl^);) In 500 mL of 6.IN
          nitfic acid.  1 mL •  10.4 mg  lead ion.                         ~

-------
    5.7  Methyl red, CAS Reg 493-52-7.  O.lt w/v.   Dissolve 0.1 g methyl  red
         1n 100 * water.
    5.8  Nickel disk — 1.5-Inch diameter by 0.020-Inch In thickness.
         Highly polished on one side.
    S.9  Nitric add, CAS Reg 7697-37-2.  (HNCh) 69-71* w/w.
    5.10 Perchloric add, CAS Reg 7601-90-3, (HC104) 70-72% w/w.

6.  Sample Collection, Preservation and Storage
    6.1  sample Collection
         6.1.1  Sampling should conform to ASTM 03370-76,  "Standard
                Practices for Sampling Water."
    6.2  Preservation and Storage
         6.2.1  Sample should be preserved by the addition of nitric add to
                a strength of 0.1N  (6 ml 69-72% per liter  of  sample).
         6.2.2  Preserved samples'need not be refrigerated for storage.
                However, analysis should be performed as soon as possible
                and  before 60 days  after collection.
         6.2.3  All  samples  shall be collected and stored  In  plastic
                containers.

7.  Calibration and Standardization .
    771Internal  proportional  counter efficiency
         7.1.1  Transfer an  aliquot containing 200-500 d/m of polonium 210
                to a 150-mL  beaker.
         7.1.2 Add 1  ml of  the lead nitrate  solution, 20 nt  concentrated
                nitric acid  and 1 at of  perchloric  acid.
         7.1.3 Cover with a watch  glass  and  evaporate to dense fumes of
                perchloric acid.
         7.1.4 .Remove from hot plate,  cool,  and dissolve residue  in 20 ml
                of 0.5N hydrochloric acid.
         7.1.5 Assemble plating cell  and fill with water to check  for
                 leaks.  Discard water if no leaks  are  present.
          7.1.6 Transfer solution to'plating cell  and  rinse the beaker four-
                more times with 20 ml of 0.5N HC1.
          7.1.7 Place plaxing cell  In boiling water.bath,  immerse  the  glass
                 stirrer. stir and plate for 4 hours.
          7.1.8  Remove stirrer, discard solution,  and  rinse disk with  water.
          7.1.9  Disassemble cell, remove disk, rinse with water and acetone,
                 dry and count for a period of time* to  accumulate at least
                 10,000 counts.          -.""••
          7.1.10 Calculate counter efficiency, c/m/d/m.
     7.2. Recovery Factor
          7.2.1  Spike triplicate l-l aliquots of tap water with a  known
                 quantity of Po-210 tracer (1000 d/n).
          7.2.2  Analyze these aliquots as prescribed in Section 9.0,
                 Procedure.
          7.2.3  Determine recovery factor (RF) as shown in Section 10.0,
                 Calculation.

 8.  Quality Control
     871General Requirements
          8.1.1  All  analysts using this method are required  to demonstrate
                  their ability  to use the method and to define their
                  respective  accuracy and precision criteria.

-------
     8.1.2  The minimum requirements for the  establishment  of  accuracy
            and precision criteria Is four replicate  analyses  of an
            externally prepared performance evaluation  sample..
     8.1.3  Application of this method to samples of  different matrix
            composition requires the analyst  to demonstrate Us
            successful use by the addition of a standardized spue
            solution and evaluation of the spike recovery.

8.2  Requirements In support of National Interim Primary Drinking Water
     Regulations (NIPOWR) regulations.
     8.2.1  The laboratory must be  certified.
     8.2.2  The laboratory must participate  once each year  in  an unknown
            performance  study  for polonlum-210  administered by EPA.
     8.2.3  The laboratory must participate  at  least twice  each year  1n
            EPA laboratory  Intercomparison studies for polonlum-210.
     8.2.4  To verify  Internal laboratory precision  for polonium-210,  a
            minimum of 10 percent duplicate  analyses must  be  performed.
     8.2.5  When  20 or more polonium-210 analyses  are  performed in a
            day,  a  performance standard and  a background sample must be
            measured with each 20 samples.  If less  than 20 samples are
            performed 1n a day. a performance sample and a background
             sample, must be measured along with the  samples.
      8.2.6. Quality control performance charts, or performance records,
             must be maintained.

 8.3  Requirements for Non-Regulated Activities
      8.3.1  It is recommended that the requirements criteria  specified
             for the NIPOWR be adopted  for all  study activities.
                  •r

 8.4  Acceptance Criteria
      8.4.1  Support of  NIPOWR .                            »,,...
           8,4.1.1   Analytical results must conform to control limits
                     established by EPA as "described In "Environmental
                     Radioactivity  Laboratory Intercomparlson Studies
                     Program - FY 1977," (EPA-600/4-77r001) or in
                      subsequent revisions.
            8.4.1.2    Duplicate measurements  are considered acceptable
                      when the  difference between them  is less than two
                      standard  deviations as  described  in EPA 600/4-77-001
                      or subsequent revisions.  . .           .,, w
            8.4.1.3    The performance standard measurement  will be
                      considered acceptable when the difference between
                      the observed or measured value and the true value is
                      less than two standard deviations as  described in
                      EPA-600/4-77-001 or  subsequent revisions.
  8.4.2     Support of Non-Regulated Activities
            8.4.2.1    It  is reconmended  that the following hierarchy be
                       used for the  setting of  accuracy and  precision
                       statements.    .              „-...,
                       8.4.2.1.1 Defined by Purpose of  Study
                       8.4.2.1.2 Defined by Interlaboratory Collaborative
                                Study
                       8.4.2.1.3 Defined by Intralaboratory -
                                Multi-operator Study

-------
                        8.4.2.1.4 Defined by Single Operitor  Study
              8.4.2.2   Duplicate measurements are considered acceptable
                        when the difference between then Is less  than two
                        standard deviations as described In Section
                        8.4.2.1.1 through 8.4.2.1.3.  Duplicate measurements
                        are considered acceptable when the difference
                        between them Is less than three standard  deviations
                        as described 1n Section 8.4.2.1.4.
              8.4.2.3   The performance standard measurement 1s considered
                        acceptable when the difference between the observed
                        and true value 1s less than two standard deviations
                        as described In Sections 8.4.2.1.1 through
                        8.4.2.1.3.  The performance standard measurement Is
                        considered aceptable when the difference between the
                        observed and true value 1s  less than three standard
                        deviations as described 1n  Section 8.4.2.1.4.

9.  Procedure
    971Neutralize  a  1-L  aliquot of the  acid  preserved sample to the basic
         side  of  methyl red using 6N NH^OH.
    9.2  Add 25 nt glacial  acetic ac~1d and .1 mL of  the  lead  nitrate  carrier
         solution.   Mix thoroughly.
    9.3  Heat  to  near  boiling on  a hot plate  and  precipitate the  lead by
         bubbling a  slow  stream of hydrogen  sulflde gas Into the  solution
         for 3 minutes.    '-   "                              ••""-.
    9.4  Remove the  hydrogen  sulfide  source,  and  continue boiling for  1
         minute.   Remove  from hot plate  and  allow to  cool until cool enough
         to safely handle.
    9.5  Filter the  solution  through  a 47 mm 0.45 micron cellulose
         triacetate  membrane  filter;   Discard the filtrate.
    9.6  Transfer the  filter  to a 100-mL beaker and add 20 ml  concentrated
         nitric  add and  1 ml perchloric add.
    9.7  Evaporate to  the first signs of dense fumes  of perchloric  acid.
         Remove from hot  plate.
    9.8  Assemble the  plating cell  and fill  with water to check  for leaks.
         Discard water 1f cell is watertight, or reassemble and  test 1f
         found to leak.                              -:
    9.9  Add 20 ml 0.5N HC1 to the perchloric acid residue and dissolve.
         Heat, If neceTsary.    •    -
    9.10 transfer solution to plating cell..  .
     9.11 Rinse beaker'four more times using 20 mL 0.5N HCT each  time and
          adding each rinse to the plating cell.      ""
     9.12 Immerse plating cell 1n a boiling water bath, position the electric
          stirring motor,  and stir without cavitation for 4 hours.
     9.13 Remove  stlrrer from cell and remove the plating cell from the water
          bath.
     9.14 Discard plating  solution and rinse cell two tines  using water
          rinses.
     9.15 Disassemble  the  cell,  and rinse both  sides of the  nickel disk with
          water and  acetone.*   -            .                  .
     9.16 Place in a 60°C  drying  oven for several minutes to dry.  Remove
          from oven  and cool  in  desiccator.
     9.17 Count for  100 minutes  in an  internal  proportional  counter.

-------
10.  Calculation
    10.1 Counter Efficiency  (E)  (Sec.  7.1)
         10.1.1 Compute counter  efficiency by calculating the ratio of the
                observed counts  per  minute to the added disintegration per
                minute (c/m/d/m)
                f    Observed count  rate (c/m)
                  "  Disintegration  rate (d/m)

    10.2 Recovery Factor (RF) (Section 7.2)
         10.2.1 Determine recovery factor by calculating the ratio of the
                observed counts  per /minute recovered to the added
                disintegration  rate.

                RF     Observed  (c/m) recovered
                   "  Added  disintegration rate  (d/m) x E

    10.3 Polonium Concentration
         10.3.1 Determine the concentration of the polonium-210 activity by
                dividing the observed counts per minute by the counter
                efficiency,  recovery factor, sample volume, and the
                conversion factor 2.22 d/'m/pCI
                Po 210 oCi/L    Sample observed  count rate
                       v  '     ~     E x RF x V  X 2.22

                where:
                E    « Counter Efficiency
                RF   m Recovery Factor
                V    * Volume in  liters
                2.22 = Conversion factor  (d/m)/(pCi)

11. Precision  and Accuracy
    11.1 Single operator precision
         11.1.1 The precision of  the method has  been evaluated by  the
                analysis of seven replicate samples at the 10.02 pCi/L
                concentration level.  The precision also includes  the
                counting error  which has  been  determined to be 31  at  the
                1 S.D. confidence level.   The  calculated precision has been
               . found  to be * 5.3%.
    11.2 Single. Operator Accuracy        .  .                         .
         11.2.1 The average concentration found  for the  seven replicate
                analyses was 10.44  pCi/L  and  ranged from a low of  9.83 to  a
                high of  11.11 pCi/L.
          11.2.2 Bias Is calculated  to be  a positive 4.11.

12. References    .
    Blanchard, R. I.,  et  al., Environmental  Science and Technology,  1967,  1,

    Figgins,  P.E., National Academy of  Science,  National Research  Council.
    1961.  NAS-NS  3037.      •         .

13. Appendix
    13.1 Plating  Cell  Assembly  (Figure  13.1)
          13.1.1  Drill  a  hole  in the bottom of an 8 oz. plastic nursing
                 bottle  to accept the glass stirrer.

-------
13.1.2 Using a band saw, cut off a 3/4 to 1-1nch section of the
       bottom to use as a cell cover to prevent excessive
       evaporation during the plating period.
13.1.3 Using emery cloth, remove the molded seams across the top of
       the mouth to make a flat seal with the nickel disk.
13.1.4 Cut out the rubber nipple and use the flange as the gasket.
13.1.5 Over-tightening of the closure tends to cause leakage.
13.1.6 Keeping the water level of the boiling water bath at a
       higher level than the plating solution prevents any minute
       leakage out of the cell during plating.
13.1.7 Weight may have to be added to the cover to keep the cell
       submerged.
13.1.8 It Is advantageous to fabricate a plastic or metal
       concentric ring for the water bath with just a large enough
       hole to accept and stabilize the plating cell.

-------
\
3---.=*
                               STIRRING
                                MOTOR
                            CUT OFF BOTTOM
                              USED AS
                               COVER
                               GLASS
                              STIRRER
                         INVERTED PLASTIC
                          NURSING BOTTLE "
                          NICKEL DISK
                          RUBBER GASKET


                          SCREW-TOP CLOSURE
 PLATING CELL ASSEMBLY

    FIGURE 13.1

-------