FINAL  REPORT
TC-3991-04
INTR0DIWI0N
Prepared by:

TETRA TECH, INC.
P/epared for:
l|.S. ENVIRONMENTA^^OTECTION AGENCY
Region <10 - Office of ihiget Sound
Seattle, WA

and

U.S. Army Corps ;of Engineers
Seattle District
Seattle; WA
 March, 1986
TETRA TECH, INC.
11820 Northup Way
Bellevue,'WA 98005

-------

-------
                                  CONTENTS





                                                                        Page



PURPOSE                                                                   1



SCOPE                                                                     2



APPROACH                                                                  4



FORMAT                                                                    5



CAVEATS                                                                   6
                                      11

-------
                                  FIGURES


Number

        Participants of the variable-ranking workshop held on
        31 May, 1985
                              ACKNOWLEDGEMENTS


        This chapter  was  prepared by Tetra Tech,  Inc.,  under the direction
of Dr. Scott Becker, for the U.S. Environmental Protection Agency in partial
fulfillment of  Contract  No.  68-03-1977.   Dr.  Thomas Ginn of  Tetra  Tech was
the Program Manager.  Mr.  John  Underwood  and  Dr.  John  Armstrong of U.S. EPA
were the Project Officers.
                                     111

-------
                                                                Introduction
                                                                     Purpose
                                                                  March 1986
                                  PURPOSE
     Environmental variables  in  Puget  Sound are measured by  a  wide variety
of organizations,  including  government agencies, universities,  and private
institutions.  However,  comparisons of results of different studies frequently
are  limited  because different methods  are  used  to measure  the  same  vari-
able(s).   The  ability  to  compare data  among  different  studies  is  highly
desirable for developing a comprehensive management strategy for the Sound.

     This  document (i.e., notebook)  presents  recommended  protocols  for
measuring selected environmental  variables in Puget Sound.   The objective is
to encourage most investigators conducting studies such as monitoring programs,
baseline  surveys,  and  intensive investigations  to use equivalent methods
whenever  possible.   If this  objective is  achieved,  most  data  from future
sampling programs should be comparable among studies.  It is recognized that
alternative  methods  exist for many of  the  variables  considered  in  this
document and that  those methods  may  produce data  of equal  or better quality
than do the recommended methods.   However, the criterion that data should be
comparable limited the  range of methods recommended in this document.  It is
also  recognized  that  future research  or  other  circumstances  may  require
modification or replacement  of one or  more of the recommended methods.  The
loose-leaf format  of  this  document was selected  specifically to allow such
changes to be made.

     The  recommendations  in  this document  pertain  primarily  to the method-
ological  specifications required  to measure  the  selected  environmental
variables.   Recommendations  for  study design  and  data analysis generally
were  not  included because those  considerations vary widely  depending upon
the  objectives of  individual  studies.   As  mentioned previously, the goal of
this document  is  to  ensure that  comparable  data  are  generated  by different
studies.   This does  not necessarily require  that  all  studies have the same
initial  design,  nor that  all  data  are  analyzed in  the same  manner  after
being   generated.  It  is  recommended, however, that  sample  collection and
analysis  specifications of study designs  be similar enough  to ensure that
comparable data are produced whenever possible.

     As an action  separate from  the preparation of  this document, several of
the  recommended protocols  will be required for use in government regulatory
permit  programs.   For  example,  the Puget  Sound  Dredged  Disposal  Analysis
(PSSDA) intends to specify several of the recommended protocols  as requirements
when conducting dredged material  regulatory testing  and disposal site assess-
ments.  Use of  such standardized procedures is essential for making  comparisons
to regulatory standards and  reference  conditions.

-------
                                                                Introduction
                                                                       Scope
                                                                  March 1986
                                   SCOPE
     A meeting was convened by U.S.  EPA  on 31 May, 1985 (Table  1) to determine
the priority of variables for protocol development or documentation.  Variables
were ranked on the basis of three major criteria:

     1.   The frequency with which each variable has been measured in a
          variety  of  studies  (e.g., monitoring  programs,  baseline
          surveys, intensive investigations) throughout Puget Sound

     2.   The  importance  of each  variable  for making decisions related
          to environmental problems in the Sound

     3.   The  degree  to which  a  variety of  methods has  been  used  to
          measure each variable.

     Using the criteria  listed above, eight groups of variables  were identified
as  having  the highest  priority for protocol  development  or  documentation.
They include:

     •    Station positioning considerations

     •    Conventional sediment variables

     •    Concentrations of organic compounds in sediment and tissue

     •    Concentrations of metals in sediment and tissue

     •    Benthic infaunal variables

     t    Sediment bioassays

     t    Pathological conditions in fish livers

     •    Microbiological  indicators.

Recommended  protocols for  all  of  these variables  are presented  in  later
sections of this document.   In  addition to  these eight groups of variables,
a  number of  others  were  considered  appropriate for  protocol  development.
The loose-leaf format of this document will allow additional protocols to be
included in the future.

     In addition to the recommended protocols for each group of variables, a
section on general  quality assurance/quality control  (QA/QC)  procedures is
included in this document.  That section identifies the major QA/QC concerns
that should be addressed when collecting and analyzing environmental samples
from Puget Sound.

-------
          TABLE 1.  PARTICIPANTS OF THE
          VARIABLE-RANKING WORKSHOP HELD
                 ON 31 MAY, 1985
     Name
Organization
John Armstrong
Bob Barrick
Scott Becker3
Joe Blazevich
Don Brown
Joe Cummins
Ray Dalseg
Bob Dexter
Burt Hamner
Ed Long
Dave Mitchell
Paillette Murphy
Jeff Osborn
Jerry Stober
Rich Tomlinson
Jack Word
U.S. EPA
Tetra Tech, Inc.
Tetra Tech, Inc.
U.S. EPA
NOAA
U.S. EPA
Metro
EVS Consultants
U.S. COE
NOAA
Metro
NOAA
Parametrix, Inc.
Univ. of Wash.
Metro
Evans Hamilton, Inc.
a Moderator.

-------
                                                                Introduction
                                                                    Approach
                                                                  March 1986
                                 APPROACH
     The recommended protocols for each group of variables were developed by
convening a  workshop comprised  of  representatives from  most  organizations
that routinely measure or use the variables  of  concern  in Puget Sound.   The
objective of each workshop was to evaluate various methods and, if possible,
agree upon which  methods  should be  used  in the  future.   Consideration  was
given to providing data that will be  comparable with the historical database.
Prior to each  workshop,  the methods  used  historically  in Puget  Sound  were
evaluated  and specific  items  requiring  standardization were  identified.
Additional  considerations  for  developing  the various recommended protocols
included data quality needs, cost, and availability of equipment and expertise.

     Each workshop  focused on  defining  acceptable methods  and determining
which of those methods would provide comparable data.   If several acceptable
methods  did  not  provide  comparable  data,  the workshop  participants  were
asked  to select  only  one  for  future use.  As  expected, a full  consensus
rarely was achieved.   However,  in  many cases,  the  majority  of participants
clearly favored a single  method.  In other instances, the participants  were
relatively evenly divided between recommending two or more methods.

     After each workshop, draft protocols were developed.   As much as possible,
recommendations of each protocol were based on the majority viewpoint of the
workshop participants.  In  some  cases, a  single recommendation could not be
given for a particular specification because no agreement was reached at the
workshop.  In  such instances, various specifications used by different Puget
Sound investigators were simply described.

     Draft  protocols  were mailed  to all  workshop participants  and  other
interested parties for written review.  Following this review, comments made
by  several  reviewers were  incorporated  into the protocols.   Most  major
comments made by  single reviewers were resolved with each respective reviewer.
After  all  written  reviews were  addressed,  protocols  were  finalized  and
included as a  chapter of this document.

-------
                                                                Introduction
                                                                      Format
                                                                  March 1986
                                   FORMAT
     Each protocol in this document is designed to stand alone.  However, in
many studies several related variables are measured simultaneously.  In such
cases, the  protocol  for one  variable may  require  some modification  to be
consistent with  that  of  a  second  variable.  For  example,  collection  of
sediment subsamples  for analysis  of conventional   variables  (e.g., particle
size, total volatile solids, total organic carbon) normally does not require
collection equipment to be washed with special  solvents.  However, if sediment
subsamples  also  will  be collected  from  the same  sample  for analysis  of
organic compounds, collection equipment for the conventional subsample must
be washed  with the  same solvent  specified  for collection  of  the organics
subsample to avoid contaminating the latter sediment.   For studies considering
multiple variables,  it  is  therefore recommended that the  protocols for all
relevant variables  be  reviewed  carefully before sampling  begins,  to ensure
that all appropriate modifications are made.

     The formats  for most  protocols  are similar  to facilitate use  of the
entire document.   The following major sections are presented for most protocols:

     •    Use  and limitations - describes what a  variable measures and
          major limitations to the use of the variable

     •    Field procedures  - describes container type, special cleaning
          procedures, collection techniques, sample quantity, preservation
          technique, storage conditions, and maximum holding time

     •    Laboratory procedures -  describes  analytical  procedures (or
          provides citations), laboratory  equipment, sources of error,
          and QA/QC specifications

     •    Data  reporting requirements  - describes  the kinds  of  data
          that the analytical laboratory should report and the units in
          which the data should be reported.

-------
                                                                Introduction
                                                                     Caveats
                                                                  March 1986
                                  CAVEATS
     Several notes of caution  require  emphasis  before  the protocols in this
document  are  presented.   First,  these protocols  were developed  solely  to
promote  the  collection of  comparable  data  in  Puget  Sound.   A  variety  of
other methods  may  exist that  produce  data of equal or  better  quality than
the  recommended  protocols.   However,  the  criterion  that  data  should  be
comparable limited the range of methods recommended in this document.

     A second caveat is that rarely was a full consensus reached with respect
to any  aspect  of any protocol.  Therefore,  it  should  not be construed that
all individuals, agencies, and institutions that participated in this effort
agreed with all of the final products.   The recommended protocols are simply
a best  effort  to represent the majority viewpoints of  the many individuals
from diverse  backgrounds  that  attended  the workshops  or commented  on  the
draft protocols.

     A  third  caveat is that  this  document is  intended  to  be dynamic.  The
loose-leaf format was selected  specifically for this  reason.  Modifications
or additions  to  the protocols can therefore  be made  in the future if needs
or viewpoints change, or if methodological refinements or improvements are made.

-------
GENQIALQA/QC

-------
FINAL REPORT
TC-3991-04                        Puget Sound Estuary Program
GENERAL  QA/QC CONSIDERATIONS
FOR COLLECTING ENVIRONMENTAL
SAMPLES IN PUGET SOUND
Prepared by:

TETRA TECH, INC.
Prepared for:

U.S. ENVIRONMENTAL PROTECTION AGENCY
Region 10 - Office of Puget Sound
Seattle, WA
March, 1986
TETRA TECH, INC.
11820 Northup Way
Bellevue, WA 98005

-------
                                 CONTENTS






                                                                        Page




LIST OF FIGURES                                                         iii



LIST OF TABLES                                                           iv



ACKNOWLEDGEMENTS                                                          v



INTRODUCTION                                                              1



SAMPLING PREPARATION                                                      2



SAMPLING PROCEDURES                                                       5



     HEALTH AND SAFETY                                                    5



     STATION LOCATION                                                     5



SAMPLE HANDLING                                                           6



     FIELD PROCEDURES                                                     6



     SAMPLE SHIPMENT                                                      7



     LABORATORY PROCEDURES                                               11



     SHIPBOARD LABORATORY ANALYSES                                       12



REFERENCES                                                               14
                                      11

-------
                                  FIGURES


Number                                                                  Page

   1    An example of a chain-of-custody record                           8

   2    Examples of a sample analysis request form and a custody
        seal                                                             9
                                  TABLES


Number                                                                  Page

   1    Examples of problems frequently encountered during offshore
        surveys and recommended solutions to each problem                 4
                              ACKNOWLEDGMENTS


     This  chapter was  prepared  originally  by  Tetra Tech,  Inc.,  for  the
Marine Operations  Division, Office  of Marine  and Estuarine  Protection,
U.S. EPA, Washington, DC as  part of U.S. EPA Contract No. 68-01-6938, Allison
J. Duryee, Project Officer.   It was prepared under  the direction of Dr.  Thomas
Ginn (Program Manager) and Dr.Scott Becker (Project Manager)  of Tetra Tech.
                                     Ill

-------
                                                               General QA/QC
                                                                Introduction
                                                                  March 1986
                                INTRODUCTION
     Quality assurance/quality  control  (QA/QC)  procedures are  necessary  to
ensure that  environmental  data achieve an  acceptable level  of  quality and
that the level of quality attained is documented adequately.   Detailed QA/QC
procedures for  the  environmental  variables considered in  this  document are
specified in each  respective recommended  protocol.   This  section describes
many of  the more  general  QA/QC  procedures  that should be  considered  when
collecting and analyzing environmental samples.   These procedures are termed
general because they pertain to samples collected for most kinds of environ-
mental  variables.   They  include  procedures  that  should  be followed  when
samples are  collected  in the field, shipped  to  laboratories,  and stored  or
distributed within laboratories.

     No workshop was  held to evaluate the  general  QA/QC procedures for use
in  Puget  Sound.  However,  they have  been  reviewed  for use  in U.S. EPA's
301(h)  marine monitoring  programs  by  U.S. EPA's  Office of  Research and
Development (Narragansett, RI and  Newport,  OR), as well as by representatives
from U.S. EPA Regions IX and X.

-------
                                                               General  QA/QC
                                                        Sampling  Preparation
                                                                  March 1986
                           SAMPLING PREPARATION
     The chief  scientist  or a designee  should  thoroughly review  the  field
survey plan (including QA/QC criteria) before each  cruise.  The  plan  should
be  checked  for  completeness  and clarity  of objectives.   A  complete  plan
should contain the following major elements:

     t    Identification of  scientific party and  the  responsibilities
          of each member

     •    Statement and prioritization of study  objectives

     •    Description of survey  area,  including background  information
          and station locations

     t    Identification of variables  to be  measured and corresponding
          required containers and preservatives

     t    Identification of all sample splits or performance samples to
          be submitted with the survey samples

     •    Brief  description  of sampling  methods, including  station
          positioning technique, sampling devices, replication,  and any
          special considerations

     •    Detailed cruise  schedule,  including time, date, and location
          of embarkation and debarkation

     •    Storage and shipping procedures

     •    Identification  of  onshore  laboratories  to  which  samples
          should be shipped after cruise completion

     •    Survey vessel  requirements  (e.g.,  size, laboratory  needs,
          sample storage needs)

     0    Location and availability of an alternate survey vessel

     •    All  special  equipment needed  for  the  survey  (e.g.,  camera,
          nets,  communication devices).

     Study  objectives  and  their prioritization should  be understood  by all
members of  the scientific  party.  This will  ensure that if modifications of
the  survey  plan  become necessary in the  field,  their  impact  on the overall
goals of  the  cruise  can be  evaluated  adequately.  After the  sampling plan
has  been reviewed, contingency plans should be outlined.  These plans should
include potential problems and  their  solutions.   Possible solutions to some

-------
                                                               General QA/QC
                                                        Sampling Preparation
                                                                  March 1986

problems frequently encountered during offshore surveys  are  listed  in Table
1.  Development  of contingency plans  can  be greatly assisted  by reviewing
cruise summary reports and consulting with chief scientists of previous cruises.

     The captain of the  survey  vessel  should be  provided with a copy of the
survey plan  to  ensure that  it  is consistent with  the  equipment and capa-
bilities of  the vessel.   Modifications  to  the ship  or cruise plan  may  be
required.

     To ensure that all required sampling equipment and supplies are available
at  the time of  sampling,  an  equipment checklist  should be  constructed.
Spare parts and backup supplies should be included in the inventory.

-------
       TABLE  1.   EXAMPLES  OF  PROBLEMS  FREQUENTLY ENCOUNTERED  DURING
         OFFSHORE  SURVEYS AND  RECOMMENDED SOLUTIONS TO EACH PROBLEM
                  Problem
          Recommended  Solution(s)
Sampling  equipment falls  or  1s lost overboard
Sampling  efficiency reduced by poor weather
Sampling  efficiency reduced because of seasickness
Sampling  delayed because vessel is Inoperative
A sample 1s partially lost or slightly contam-
inated

Bottom sampler will not penetrate  to a sufficient
depth after repeated  casts
One  or more water bottles fail to  trip  on
a particular cast
- Have  necessary tools  onboard to make  repairs
• Maintain spare  parts  Inventory for major
  equipment
- Have  back-up equipment onboard
- Have SCUBA equipment  and divers onboard
  for retrieval
- Know  where nearest tools, back-up equipment.
  or divers are located on shore

- Extend  cruise length
- Reschedule cruise

- Ensure scientific crew is large enough
  to compensate for reduced personnel
- See solutions to second  problem

- Reschedule cruise
- Extend  cruise length  once vessel Is repaired
- Charter an alternate  vessel

- Discard and take another sample
  Add  weight to the sampler
  Move a short  distance before  taking next
  sample

  Discard all  samples from  that  cast and
  take a  new cast

-------
                                                               General  QA/QC
                                                         Sampling Procedures
                                                                  March 1986
                            SAMPLING PROCEDURES
HEALTH AND SAFETY

     Guarding the health and  safety of the sampling team not only is mandatory
in itself,  but  also ensures that  such  concerns do  not interfere with  the
collection  of  quality  data.   It  is recommended  that all  members  of  the
sampling  team  wear hard  hats  and Coast  Guard-approved personal  flotation
devices when working  on deck.   Whenever water  sampling, grab  sampling,  and
trawling are being  conducted,  the danger exists  that  the  block,  winch,  or
hydrowire will  fail  or  that  the water  sampler, grab,  or trawl  assembly
(especially the otter boards) will swing  out  of control.   Any one of  these
events could knock someone unconscious (or worse), and possibly overboard.

     When sampling in  areas that are  contaminated by toxic materials,  special
equipment (e.g., detectors, respirators) may be required by federal or  state
regulations.  The use of  plastic  safety glasses when working with sediments
is a  good  practice.   Sediment particles  introduced  to the eye  can  cause
abrasion, infection, or chemical contamination.

     Perhaps the  most common ailment when  sampling  is "mal de  mer"  (i.e.,
seasickness).  Although much  humor has evolved  at the expense of the seasick,
sample quality  can  be seriously compromised by impaired deck  hands.   It is
therefore recommended  that  members  of  the  sampling  team  who  are prone  to
seasickness take appropriate medication (e.g., dramamine,  scopolamine)  prior
to the cruise.

STATION LOCATION

     Accurate  navigation is  essential  to  ensuring  that  stations  can  be
plotted and reoccupied with a high degree of  certainty. Although a  variety
of navigation and/or position fixing  systems are available currently, factors
such as price, availability,  and accuracy vary considerably among them.  The
station  positioning system selected  for  a given  survey  should  be  able to
meet all  study  design requirements  for accuracy  and should, at  a minimum,
provide a high degree of precision (i.e., repeatablemeasurements).  Positioning
systems  that  are precise but  lack  a high  degree of  accuracy may be used
after actual station locations are determined by accurate,  independent  means
(i.e.,  "ground-truthed").  For bottom-related  samples,  all  positioning
systems  should  be used  in  conjunction  with  a  fathometer  to  ensure that
sampling occurs at  the  proper water  depth  (allowing  for tidal  stage  and any
fathometer corrections).  The protocols for station positioning presented in
a later chapter of  this  notebook  should be  consulted for details related to
locating stations.

-------
                                                               General  QA/QC
                                                             Sample Handling
                                                                  March 1986
                              SAMPLE HANDLING


     After sample collection, proper sample handling ensures that changes in
the constituents  of  interest are minimized  and  guards against  errors  when
shipping  and  analyzing samples.  All  stages of  sample handling  should  be
documented adequately.  Documentation helps  ensure  that all  sample handling
requirements are carried out and serves as proof that handling was conducted
properly if questions arise later.

FIELD PROCEDURES

     It is important  throughout  any  sampling  and analysis program to maintain
integrity of  the sample from the time  of collection  to  the point  of  data
reporting.    Proper  chain-of-custody  procedures allow the  possession  and
handling  of  samples   to be  traced  from  collection  to final  disposition.
Documents needed to maintain proper chain-of-custody include:

     •    Field logbook —  All pertinent  information  on  field activities
          and sampling  efforts  should  be recorded in  a bound logbook.
          The field  supervisor  should be responsible for ensuring that
          sufficient  detail  is  recorded  in  the  logbook.   The  logbook
          should enable  someone else  to  completely  reconstruct the
          field  activity  without relying  on the memory  of  the  field
          crew.  All  entires should  be made in indelible ink, with each
          page signed and dated by the author, and a line drawn through
          the remainder of any page.  All corrections should consist of
          permanent line-out deletions that are initialed.  At a minimum,
          entries in a  logbook should include:

               Date and time of starting work

               Names of field supervisor and team members

               Purpose  of proposed sampling effort

               Description  of  sampling  site,  including information on
               any photographs that may  be taken

               Location of sampling site

               Details of actual sampling effort, particularly deviations
               from standard operating procedures

               Field  observations

               Field measurements made (e.g., pH, temperature, flow)

-------
                                                              General QA/QC
                                                            Sample Handling
                                                                 March 1986
               Field laboratory  analytical  results

               Sample identification

               Type and  number of  sample  bottles  collected

               Sample  handling,  packaging,  labeling,  and  shipping
               information  (including  destination).

          Chain-of-custody procedures  should  be maintained  with  the
          field logbook.  While being used in the  field,  the  logbook
          should remain with the  field  team at all times.  Upon completion
          of the  sampling  effort, the logbook  should  be  kept  in  a
          secure area.

     •    Sample labels  —  Sample  labels  must be  waterproof  and must be
          securely fastened to  the  outside  and/or  placed  inside  each
          sample container  (depending  on  the kind of sample) to prevent
          misidentification of samples.  Labels  must contain  at  least
          the sample number,  preservation  technique, date and  time of
          collection, location  of collection, and   signature  of  the
          collector.  Labels   should be  marked with  indelible  ink.
          Abbreviated labels  may also  be placed on the cap of  each  jar
          to facilitate sample identification.

     •    Chain-of-custody records — A chain-of-custody record (Figure 1)
          must accompany every sample.  Each person  who has custody of
          the sample must sign  the form and ensure that  the  samples  are
          not left unattended unless  secured properly.

     •    Custody  seals — Custody seals (Figure 2)  are used  to detect
          unauthorized tampering with  the  samples.   Sampling  personnel
          should  attach  seals  to all  shipping  containers  sent  to  the
          laboratory by common carrier.  Gummed paper seals or custody
          tape should be used so  that the  seal must  be  broken when  the
          container holding the samples is  opened.

For further information  regarding  proper chain-of-custody procedures, consult
the policies and procedures manual  for  the National Enforcement Investigations
Center (NEIC; U.S. EPA 1978).

SAMPLE SHIPMENT

     All preseved  samples  should  be shipped immediately  after completion  of
sampling.  This minimizes the number of people handling samples, and protects
sample quality  and  security.  Guidance for shipping hazardous materials  can
be found in U.S. Department of Transportation (1984).   As samples are prepared
for shipping, the following should be kept in mind:

-------
&ER&
i
E
/
(




Jmted Slant Ragi
kgencv Sett
:HAIN OF CUSTODY RECORD
on 10
Sixth Avenue
tie WA 98101


PROJECT
LAB #











STATION











DATE











TIME













RELINQUISHED BY: «*»«**
RELINQUISHED BY: 'Svwu*
RELINQUISHED BY: rsgrwu*
RELINQUISHED BY: Mgmu«
DISPATCHED BY: u«nm«* DATEi
METHOD OF SHIPMENT:





SAMPLERS: is****
SAMPLE TYPE
i
3











M
m
Z
-i











TISSUE











&











9











o
I











~* O
> f*l
2 3D
a
M











REMARKS











RECEIVED BY: «w»w
RECEIVED BY: «*.««
RECEIVED BY: a*wu*
RECVD BY MOBILE LAB FOR FIELD
ANAL.: ttpwu*
'TIME
RECEIVED FOR LAB BY: <**»*>*

DATE;
DATE/
DATE
DATE
DATE
TIME
'TIME
mME
mME
mME




Distribution Onginil- Accompany Shipmtni
On* Coov- Suray CoorAnttor FWd Fita
Figure 1. An example of
a chain-of -custody record.


8

-------
U.S. ENVIRONMENTAL PROTECTION AGENCY
CLP Sample Management Office
P.O. Box 818 - Alexandria, Virginia 22313
Phone: 703/JJ7-2K90  - FTS/JJ7-2490
                                   SAS Number
                          SPECIAL ANALYTICAL SERVICE
                                 PACKING LIST
Sampling Office:
Sampling Contact:
(name]
(phone)
Sampling DateUh
Date Shipped:
Site Name/Code;

Ship To:
Attn:
For Lab Use Only
Date Samples Rec'd:
Received By:

       Sample
      Numbers
   Sample Description
Analysts, Matrix, Concentration
Sample Condition on
  Receipt at Lab
 I.
 2.
 3.
 ».
 5.
 6. .
 7.
 >• .
 9.
10. .
M.
12.
13.
14.
19.
16.
17.
IS.
19.
20.
                                                                For Lab Uw Only
                             EPA CUSTODY SEAL

ENVIRONMENTAL PROTECTION AGENCY

•AMPLB HOb
UQNATUNB
PMMT HAW AND TITLB (M^MM* ****** «v Tm*mt*mV
•
j
3
•

1

    Figure 2.   Examples  of  a sample analysis request form  (above)
                 and a custody seal  (below).

-------
                                                          General  QA/QC
                                                        Sample Handling
                                                             March 1986

•    Shipping containers  should  be in  good  shape and  capable  of
     withstanding rough treatment during shipping.

•    Samples should be packed tightly:

          Dividers must separate all glass containers

          Empty space  within shipping  boxes  should  be filled  so that
          jars are held securely.

•    All  containers  must  be leak-proof.  If  a container  is  not
     leak-proof by design,  the  interior should be lined  with  two
     heavy-duty plastic  bags  and the  tops of  bags should be tied
     once samples are  inside.  Adequate  absorbent  material  should
     be placed in the container in a quantity sufficient to absorb
     all of the liquid.

•    All samples should be accompanied by a sample analysis request.
     Variables to be analyzed by  the laboratory,  and total  number
     and kind of  samples shipped  for analysis  should be listed on
     the request  sheet.  An example sample analysis  request form
     is illustrated in Figure 2.   The laboratory should acknowledge
     receipt  of  shipment  by signing and  dating  the  form,  and
     returning a copy to the designated QA coordinator.

•    A chain-of-custody  record  for  each shipping  container should
     be filled out completely and signed.

•    The  original  chain-of-custody record  and  analysis request
     should be protected  from damage and  placed inside the shipping
     box.   A  copy of each  should be  retained  by   the  shipping
     party.

•    The custody  seal  should  be  attached so  that  the shipping box
     cannot be opened without breaking the seal.

•    For shipping containers carrying liquid samples:

          A "This End Up" label should be attached to each side to
          ensure that jars are transported in an upright position.

          A "Fragile-Glass" label should be attached  to the top of
          box to minimize agitation of samples.

•    Shipping  containers should  be sent  by  a carrier  that will
     provide  a delivery  receipt.   This  will  confirm that  the
     contract  laboratory  received the samples  and serve  as a
     backup to the chain-of-custody record.
                                 10

-------
                                                               General  QA/QC
                                                             Sample Handling
                                                                  March 1986

     •    All shipping charges should be prepaid by the sender to avoid
          confusion and possible  rejection  of package by  the contract
          laboratory.

LABORATORY PROCEDURES

     At the  laboratory,  one  person  should be  designated  custodian of  all
incoming samples.  An  alternate  should also  be  designated to serve in  the
custodian's absence.   The custodian should oversee the following activities:

     •    Reception of samples

     a    Maintenance of chain-of-custody records

     •    Maintenance of sample tracking logs

     •    Distribution of samples for laboratory analyses

     •    Sending of samples to outside laboratories

     •    Supervision  of  labeling,  log  keeping,  data  reduction,  and
          data transcription

     •    Storage and security of all samples, data, and documents.

     Upon reception of  samples,  a designated  laboratory person  should fill
out  the  chain-of-custody  record,  indicating time  and  date  of reception,
number of samples, and condition  of  samples.  All  irregularities indicating
that sample security or quality may have been  jeopardized  (e.g., evidence of
tampering, loose lids, cracked jars) should be  noted on  the sample analysis
request form and  returned  to  the client-designated QA  coordinator.   In
addition, a  designated laboratory person  should initiate  and  maintain  the
sample -tracking  log  that  will  follow  each  sample  through  all  stages  of
laboratory processing and analysis.

     Minimum information  required in  a sample  tracking  log  includes  the
following:

     •    Sample identification number

     •    Location and condition of storage

     •    Date and time of each removal of and return to storage

     •    Signature of person removing and returning the sample

     •    Reason for removal from  storage

     •    Final disposition of sample.


                                      11

-------
                                                               General QA/QC
                                                             Sample Handling
                                                                  March 1986

     All logbooks,  labels,  data  sheets,  tracking logs, and  custody  records
should have proper identification numbers and be accurately filled out.  All
information should be written in ink.  Corrections should be made by drawing
a line through the  error  and  entering  the correct information.  Corrections
should be  initialed  and  dated.   Accuracy  of  all data reductions  and tran-
scriptions should  be verified  at  least  twice.   All  samples  and  documents
should be properly  stored within the laboratory  until  the client authorizes
their removal.   Security  and confidentiality of  all  stored material should
be maintained at all times.  Before releasing analytical  results, all infor-
mation  on  sample  tags,  data  sheets,  tracking  logs,  and  custody  records
should be  cross-checked  to ensure that  data  pertaining  to  each  sample are
consistent throughout the record.

     Originals of the following documents should be sent to the client:

     •    Chain-of-custody records

     •    Sample tracking logs

     •    Data report sheets

     •    Quality control records.

Copies  of  all  forms should be  retained  by the laboratory  in case originals
are lost in transit.

SHIPBOARD LABORATORY ANALYSES

     Depending  upon the  size and capabilities  of the  survey vessel, many
environmental variables can be analyzed on board.  In general,  the  laboratory
procedures described for the protocols  in this  document  are applicable to
both shipboard  and  land-based laboratories.  This  consistency is important
to  ensuring  that analytical results will  be  comparable  regardless of which
kind of laboratory  generates  them.

     Although  most  laboratory procedures  are similar between shipboard and
land-based laboratories,  a number of additional  factors  must be  considered
when analyzing samples at sea.  These factors relate primarily to the remoteness
of  the shipboard  laboratory  from land-based  support and  the'movement and
limited space of the survey vessel.  The major considerations are as  follows:

     •    The design of the laboratory should be efficient, with convenient
           equipment locations and  adequate storage  space.

     •    The  vessel should  be  equipped with an uninterruptible  power
           supply  that  is  adequate  for  operation  of  all  scientific
           instruments.
                                      12

-------
                                                     General  QA/QC
                                                   Sample Handling
                                                        March 1986

The laboratory should be well  ventilated  to  remove any  toxic
vapors created by chemicals.

The  temperature of  the  laboratory  should be  controlled,
especially if variations in ambient temperature  can influence
particular analyses.

Adequate lighting  is  necessary,  especially  for  analyses
requiring color discrimination (e.g., titration  endpoints).
The laboratory should have adequate water purification apparatus
or be capable of storing water purified on shore.

For storing many kinds of  samples, adequate refrigeration and
freezing capabilities are  desirable.

The laboratory  should never be used as a general  passageway
or lounge.

The laboratory should be off limits  to unauthorized personnel.

Adequate  safety  and first  aid equipment  should be on  board,
preferably including an overhead quick-pull safety shower.

Extreme care must be taken when handling samples (for quality
purposes) and hazardous reagents  (for  safety  purposes),  as
vessel movement can sometimes be unpredictable.

Backup  supplies  and  instruments  should  be on  board  so that
sampling  can  continue if  a  piece  of equipment  is  broken or
will not  operate properly.   A continuously updated inventory
tracking system is useful  for maintaining backup equipment.

All equipment should be  properly secured to compensate for
predictable  and  unpredictable  vessel  movements.    Specially
designed racks are useful  for this purpose.

Instruments should  be checked and calibrated before sailing,
so  that problems  requiring  land-based  assistance  can  be
solved quickly.

Whenever  possible,  plastic  containers  should be used instead
of glass because plastic is less susceptible to  breakage.

Instruments having  digital  displays are preferred over those
using needles.

Pre-printed  data sheets  should be  used  to ensure  that all
required information  is recorded.
                            13

-------
                                                               General  QA/QC
                                                                  References
                                                                  March 1986
                                REFERENCES
U.S. Department of Transportation.  1984.  Hazardous materials  regulations.
U.S. DOT, Washington, DC.   Federal  Register Vol. 49, Chapter 1,  Subchapter
C.  pp. 52-792.

U.S. Environmental Protection Agency.   1978  (revised 1983).  NEIC  policies
and procedures.   EPA-330/9-78-001-R.   National  Enforcement Investigations
Center, Denver, CO.
                                      14

-------
STATION POSmONING

-------
FINAL REPORT                       Pygf|J?ounff Estuary Program
TC-3090-05
RECOMMENDED PROTOCOLS
FOR STATION POSITIONING
IN PUGET SOUND
Prepared by:
TETRA TECH, INC.

Prepared for:
RESOURCE PLANNING ASSOCIATES

for:
PUGET SOUND DREDGED DISPOSAL
ANALYSIS (PSDDA)
Monitored by:
U.S. Army Corps of Engineers
Seattle District
Seattle, WA
August, 1986
TETRA TECH, INC.
11820 Northup Way
Bellevue, WA  98005

-------
                                 CONTENTS

                                                                        Page
LIST OF FIGURES                                                          iv
LIST OF TABLES                                                           vi
ACKNOWLEDGMENTS                                                         vi i
INTRODUCTION                                                              1
POSITIONING METHOD SELECTION PROCEDURE                                    3
    DEVELOPMENT OF SAMPLING PROGRAM                                       3
       Physical Conditions at the Study Site                              5
       Equipment and Analyses                                             5
       Station Separation                                                 6
       Reoccupation                                                       6
       Program-Imposed Constraints                                        7
    DEFINITION OF POSITIONING REQUIREMENTS                                7
    REVIEW AND SELECTION OF POSITIONING METHOD                           10
    IMPLEMENTATION OF POSITIONING METHOD                                 11
RECORDKEEPING REQUIREMENTS                                               12
    FIELD RECORDS                                                        12
       Initial Survey Description                                        12
       Day Log Entries                                                   12
       Station Log Entries                                               13
    REPORTING REQUIREMENTS                                               14
RECOMMENDED POSITIONING ACCURACIES                                       15
    CLASSIFICATION OF SAMPLING                                           15
    RECOMMENDED ACCURACY                                                 15
    RECOMMENDED POSITIONING METHODS                                      17
REFERENCES                                                               22
                                     ii

-------
APPENDICES

APPENDIX A:  SITE-RELATED POSITIONING LIMITATIONS                       A-l

APPENDIX B:  POSITION ERROR ANALYSIS                                    B-l

APPENDIX C:  POSITIONING METHODS AND CONSIDERATIONS FOR SAMPLING IN
          PUGET SOUND                                                   C-l

    OPTICAL POSITIONING TECHNIQUES                                      C-l

    ELECTRONIC POSITIONING TECHNIQUES .                                 C-14

    RANGE-AZIMUTH SYSTEMS                                              C-43

    REFERENCES                                                         C-49

APPENDIX D:  EVALUATION OF POSITIONING METHODS                          D-l

    POSITIONING PROCEDURES IN USE IN PUGET SOUND                        D-l

    CANDIDATE SYSTEM OVERVIEW                                           D-4

    SCREENING CRITERIA                                                  D-6
                                     m

-------
                                  FIGURES

Number                                                                  Page
   1    Positioning method selection procedure                            4
   2    Components of error in probable area of sample collection         9
   3    Comparison of approximate accuracy ranges of some positioning
        methods                                                          20
APPENDIX A
  A-l   Regions of Loran-C signal interference in Puget Sound           A-3
APPENDIX B
  B-l   Line of position measurements for two shore stations
        depicting LOP uncertainty (a), and associated error
        indicator (b)                                                   B-2
  B-2   Angle-of-cut effects on fix accuracy                            B-4
  B-3   Illustration of radial error                                    B-7
  B-4   Components of error in probable area of sample collection      B-10
  B-5   Effects of wire angle on the probable (P=0.68) area of
        sample collection                                              B-ll
  B-6   Effects of navigational accuracy on the probable (P=0.68)
        area of sample collection                                      B-13
APPENDIX C
  C-l   Station positioning by theodolite intersection                  C-4
  C-2   Station fix using position circles                              C-8
  C-3   Three-arm protractor  for sextant resections                     C-9
  C-4   Shore target locations to avoid the danger circle              C-ll
  C-5   Operating modes for radio navigation systems                   C-19
  C-6   Seasonal variations in Loran-C signals at Neah Bay,
        Washington 5990                                                C-32
                                     IV

-------
  C-7   Regions of Loran-C signal interference in Puget Sound          C-35
  C-8   Range-azimuth positioning system area of coverage              C-44
APPENDIX D
  D-l   Approximate accuracy versus distance of some positioning
        methods                                                         D-8

-------
                                  TABLES

Number                                                                  Page
   1    Contributors to the station  positioning protocols                 2
   2    Classification of common types of sampling                       16
   3    Summary of acceptable positioning methods                        18
APPENDIX B
 B-l    Probability versus R/o for elliptical  bivariate distributions
        with two equal standard deviations                              B-5
 B-2    Circular error probabilities as a function of measurement
        standard deviation ratios and error circle radius               B-8
APPENDIX C
 C-l    Summary of vessel positioning methods                            C-2
 C-2    Summary of vernier transit and scale-reading theodolite
        characteristics                                                 C-6
 C-3    Summary of micrometer and digitized theodolite characteristics  C-7
 C-4    Marine sextant characteristics                                 C-13
 C-5    Variable Range Radar (VRR) system characteristics              C-15
 C-6    Electronic positioning system categories                       C-17
 C-7    Electronic distance measuring instruments                      C-21
 C-8    Total station characteristics                                  C-23
 C-9    Short-range positioning system characteristics                 C-25
 C-10   Medium-range positioning system characteristics                C-29
 C-ll   Long-range positioning system characteristics                  C-30
 C-12   Range-azimuth positioning system characteristics               C-45
APPENDIX D
 D-l    Interviewed governmental and private groups that conduct
        sampling in Puget Sound                                         D-2
                                     vi

-------
 0-2    Characteristics of positioning methods  used  for sampling in
        Puget  Sound                                                    D-3

 D-3    Evaluation of navigation methods for station positioning        D-7

 D-4    Typical  equipment rental costs (1985 dollars)                  D-13
                             ACKNOWLEDGEMENTS


     This chapter was prepared  by Tetra Tech,  Inc., under  direction of
Mr. Jeff Stern,  for Resource Planning Associates  (RPA)  for the Puget  Sound
Dredged Disposal  Analysis  (PSDDA), monitored  by the Seattle District, U.S. Army
Corps of Engineers (COE) in partial fulfillment of Contract No. DACW67-85-D-
0029.   Dr.  David Kendall  of the U.S.  COE  was  the Technical Coordinator,
and Mr. Gary Bigham of Tetra Tech  was  the Program  Manager.  Dr. William
Muellenhoff of Tetra  Tech coauthored  the  appendix on positioning methods
and helped resolve many of  the technical  issues that  arose as the  protocols
were being prepared.
                                    VI1

-------
                                                       Station  Positioning
                                                              Introduction
                                                               August  1986
                               INTRODUCTION
     This chapter  presents recommended  procedures for station positioning
during environmental sampling in Puget  Sound.  Specifically,  the objectives
of this chapter  are to:

     •    Recommend methods for selecting  station positioning methods

     •    Recommend  positioning  accuracies and  associated methods
          for specific kinds of sampling.

     The recommendations in this  chapter are based  on the results of a
workshop and written reviews by representatives of most agencies, institutions,
and private  firms that  fund  or conduct environmental research  in Puget
Sound (Table  1).  At the workshop,  it was  agreed that no single positioning
method  would be adequate  for  all sampling scenarios.  Instead, the group
recommended development  of  procedures to select appropriate positioning
methods for specific sampling objectives and to ensure consistent implementation
of those methods.  A standardized selection procedure was recommended that:

     •    Accommodates different sampling objectives and site-specific
          conditions

     •    Does  not limit the  positioning methods that can provide
          the accuracy needed or specified

     •    Allows investigators to use familiar or available systems

     •    Allows the funding  agencies to specify levels of accuracy
          appropriate for sampling objectives or database entry

     •    Incorporates positioning accuracy as a component of sampling
          study  design regardless of scientific approach.

     The description  of the selection  procedure begins with a  review of
interrelationships  between sampling program design  and positioning  require-
ments.   Guidance for defining project-specific  requirements is  provided.
The procedures and  equipment that can provide  the  required  accuracies  are
identified and  an evaluation procedure  for comparing candidate positioning
methods is presented.  Requirements  to standardize the  implementation of
positioning  methods and reporting of  station locations are also presented.
Finally, accuracy levels and corresponding  positioning methods are recommended
for three categories of sampling activities (areal samples, point  samples,
and special studies).

-------
   TABLE 1.  CONTRIBUTORS TO THE STATION POSITIONING  PROTOCOLS
Name                                    Organization

John Armstrong*                         U.S.  EPA
Gary Bighama»B                          Tetra Tech,  Inc.
John Oermodya                           Raven Systems
Ben Huntley3                            Sea-Lease,  Inc.
David Jamison3                          WDNR
Karl Kassebaum                          U.S.  EPA
David Kendall3                          U.S.  COE
Gary Mausetha                           Nortec
Gary Minton3                            Resource Planning  Assoc.
Bill Muellenhoff3                       Tetra Tech,  Inc.
Bob Parker                              U.S.  COE
Tony Petrillo3                          Nortec
Keith Phillips3                         U.S.  COE
Anthony Roth3                           Cooper Consultants
Debra Simecek-Beatty3                   NOAA
Jeff Sterna»b                           Tetra Tech,  Inc.
Steve Til ley3                           WDNR
Bruce Titus3                            URS Engineers
Frank Urabeck                           U.S.  COE
Pete Wilkinson3                         Battelle Northwest
3 Attended the workshop held on January 27, 1986.

b Workshop moderators.

-------
                                                       Station  Positioning
                                                          Method Selection
                                                               August 1986
                  POSITIONING METHOD SELECTION PROCEDURE
     A procedure  for identifying  thfe  positioning accuracy appropriate for
a specific sampling objective and the  corresponding positioning  methods
is presented  in this section.  The procedure involves identifying  specific
factors of the  sampling program that affect  the  accuracy and  feasibility
of positioning methods.  Next,  the individual  stations of the  sampling
program are used to determine the specific  positioning requirements.  Available
systems are then evaluated based on  positioning requirements and site-specific
constraints to  select an appropriate positioning  method.  These  steps are
followed  by approval and procurement  of the selected system,  assembly and
installation of shore stations (if needed), and  equipment calibration and
field-testing.   The procedure is presented graphically in Figure 1.

DEVELOPMENT OF  SAMPLING PROGRAM

     Study design  and  location  are  important factors  in  the evaluation
of positioning  methods for environmental sampling vessels.   Study design
affects  required positioning accuracy and location affects feasibility
and accuracy of some  positioning methods.   Specific design  and location
factors include physical conditions  and  topography of the study site, proposed
equipment and analyses, minimum station  separation, station  reoccupation,
and program-imposed constraints.  All  of these factors need to be  considered
prior to sampling to ensure that positioning  methods do not compromise
quality  or interpretation  of the data.  The design factor most sensitive
to station location  (e.g.,  requiring  the highest positioning  accuracy)
will  initially define the positioning  requirements.  This initial  estimate
should be re-evaluated during the study design to determine whether positioning
limitations will require changes in  the  sampling program.

     The type  of  positioning  accuracy that is needed to meet the  specific
sampling objectives of each study should be identified during the  evaluation
of effects of  design and location factors on  positioning  requirements.
Absolute or predictable accuracy refers  to a method's ability  to  correctly
define  a  position by latitude and longitude (Bowditch 1984).   Repeatable
or relative accuracy measures a method's ability to return  the  user to
the same position  time after time.  The  difference between these two accuracies
can be significant.   Some of the study design or location factors  may  affect
one  type  of accuracy but not the other.  Examples are presented  under each
subheading below.   It is important to identify which accuracy is of  concern
for  positioning during a specific sampling  program before finishing the
following evaluation  of effects on study design.

-------
PURCHASE
OR LEASE
SYSTEM


/FUNDING /
AUTHORIZATION /
Figure 1.   Positioning method selection procedure.

-------
                                                       Station Positioning
                                                          Method Selection
                                                              August 1986
Physical  Conditions at the Study Site
     The ability  of  a  positioning method to achieve its  highest projected
accuracy depends,  in part, on site-specific  conditions.  The accuracy of
a position  fix  from any two points  increases as the angle between the lines-
of-position (LOP)  on which the vessel must lie approaches 90 degrees.   The
area of  probable location that can be  resolved by a navigation method  (the
radial error) varies among locations because of changes in the  relative  location
of the fix  points  and the vessel.  The number of objects or targets with  known
locations that  can be used for a position fix  depends on the methods  used
and, in  some  cases, on  the  surrounding terrain.  An acceptable fix target
for one method may not be  an acceptable target for another method.  A preferred
method may not be usable or sufficiently accurate at all  locations.  For
example, Loran-C cannot be used in some parts of Puget Sound  (see Figure A-l,
Appendix A)  and the accuracy of  visual  sighting  methods decreases with
distance from  shore.  Thus, the location (or the combination of locations)
of  the  study  is a principal  determinant in the usefulness of a specified
positioning method.  Effects of location on the accuracy and applicability
of  various positioning  methods are  described  in  Appendix A for  optical,
radar range, and short-range and long-range electronic positioning methods,
which are commonly used in Puget Sound.

     Weather,  currents,  and other physical  factors  may  also  reduce the
achievable  accuracy of  a  positioning method.   For example, the  relative
drift of the  sampling equipment  away from the boat under strong  currents
or winds can increase with depth.  Resulting  positioning errors  in sample
location (as  opposed  to boat location) may exceed acceptable limits for
the study if effects of site location on positioning accuracy are not considered
during design  of the sampling program.

Equipment and  Analyses

     The vessel  available for  the sampling program  may  affect the  study
design by limiting the  types of positioning equipment that can be  deployed
or  set  up  on  board.   If a positioning  system already exists on the vessel,
it  should be evaluated  to determine whether  its  accuracy  is adequate for
the sampling program.

     Different levels of  accuracy are  required for different  kinds of  sampling.
Water column sampling generally does not require a precisely known  station
location because the  water column  is  relatively homogeneous  compared with
sediments.   Trawling transects do not  require high positioning accuracies
because  the sampled area  is large and because the precise location of the
net at any  specified moment is uncertain.  Accuracy is  much  more  important
for sampling conducted  with equipment  that penetrates or rests  on  the bottom
(e.g., cores,  grabs).   Heavier equipment will  usually  reduce wire angles
(see discussion  under "Define Positioning  Requirements," below) and the
area in which  the sampler was  probably  located.  For equipment  left at
one location  for a period  of time  (e.g., current meters, sediment traps),

-------
                                                       Station  Positioning
                                                          Method  Selection
                                                               August 1986
repeatable  or  relative accuracy to return to the same position  is  important
for retrieval.  Absolute or exact geodetic positions may be less  critical.
Sampling  of point sources  generally requires both high  absolute accuracy
for exact location  of sources and high relative accuracy for  proper definition
of the spatial  distribution of sediment pollutant concentrations.

     The chemical  or statistical  analyses to which the  collected  samples
are subjected should also be considered in determining the required  navigational
accuracy.   If  a  gradient of  environmental effects is suspected,  but the
analytical  technique  cannot  measure  small  differences in the value of a
specified  variable,  sampling stations can be  located farther  apart and
a relatively less accurate positioning method can be used.   However,  within-
station  variability  may be more difficult  to discern using a less accurate
positioning  method.   For variables with a patchy distribution,  the patch
size  could  be  smaller than the area defined by  the repeatable accuracy
of the positioning method, resulting in replicates sampled  across  community
or  physical boundaries.  These conditions may not be noticed  in  the  field
and could prevent correct interpretation of the data.  Statistical  comparisons
with replicate  samples (e.g., "synoptic" data, field replicates, time-series
samples)  from heterogeneous stations will deserve special  attention.  The
effects  of  navigational  positioning accuracy and the associated probable
sampling area (area from which samples could have been collected)  on statistical
comparisons  of  data should be considered in  the study design.

Station Separation

     When sampling stations occur in a grid, along a gradient,  or  in highly
heterogeneous areas,  it is especially important to maintain the  statistical
separation between them.  Minimum separation should be defined by the diameter
of the probable sampling  area.  Spatial resolution of station  locations
(i.e., probable  sampling area) is limited by positioning method  accuracy,
depth, and wire angle.  The proposed  positioning method can  be  evaluated
to determine whether  statistical separation  for the proposed station  config-
uration will  be achieved.  Probable sampling area is determined  by calculating
the positioning  error at a 95 percent probability level  and  adding the
effect of wire  angle.  Procedures for determining the spatial  resolution
of  a  single station  and  adequate station separation are presented  later
in this section under "Define Positioning Requirements."

Reoccupation

     Station reoccupation  (e.g.,  for  replicates or  time series samples)
requires  accurate positioning techniques  to  minimize the  area sampled.
The variance (s2) of the measured variables  (for a specified  number of
replicates)  usually increases with the area  sampled (Holme and Mclntyre
1984).   The increase in  variance can mask trends or gradients that may
actually exist.  With less accurate positioning  methods,  more replicates
are needed  to  properly  characterize the  region and replicate samples are
more likely to have  been collected from different patches within the probable

-------
                                                       Station  Positioning
                                                          Method  Selection
                                                               August 1986


area sampled.   Because repeatable accuracy  is typically one order of magnitude
greater than  absolute accuracy for a specified navigation method, repeatable
accuracy should be used to determine the  probable area sampled at a  reoccupied
station.  (If the exact geodetic position is  important, a  highly  accurate
system  can  be used for  initial  occupation and a less accurate method can
be used for  reoccupation.  The repeatable accuracy of the  less  accurate
method may approach the absolute accuracy of the more accurate method.)

Program-Imposed Constraints

     Various  program-specific constraints may affect the decision on positioning
methods.  Sampling time limitations may preclude more logistically  demanding
positioning  methods.  The  proximity of  the study and reference areas to
one another may require the shore  stations to  be moved.   Operator  (e.g.,
investigator, vessel captain) experience with the proposed  positioning
system can affect performance.  An experienced operator is  more  likely
to arrange  transponders  (electronic systems) or  landmarks (range or  bearing
methods) for complete coverage of the  study  area and is more likely to
identify potential problem areas prior  to the start of the survey. Contractual
obligations may limit changes to the study  design, establish maximum  expend-
itures  for  station positioning, or require a minimum achievable accuracy
for the survey. Unless the sampling design demands higher accuracy,  funding
agency specifications for positioning requirements should be observed.

DEFINITION OF POSITIONING REQUIREMENTS

     Once a  study design  is drafted, project-specific positioning requirements
can be established.  Such requirements  are  influenced primarily by:

     •    Characteristics  of the survey  area  (e.g., physical,  geo-
          graphic)

     •    Minimum required station separation

     •    Maximum acceptable area from which  the replicate  samples
          can be collected.

     Physical  and  geographic  characteristics  of  the survey  area affect
the  applicability of various positioning  methods and  their  achievable
accuracies.   For example,  as noted earlier, signal interferences  in  some
parts of Puget Sound may preclude the use  of Loran-C as a positioning  method.
Accuracy of  other methods  may be insufficient because  objects by which
to establish the position fix (shore stations or  targets)  are inaccurately
mapped  or unavailable.  Map accuracy  and  the  ability to locate on a map
the shore stations or objects from which  fixes  were taken are the  largest
potential sources of positioning error.  No matter how accurate the positioning
methods are,  the locations  of  sampling  stations will  always be relative
to the  location of fix  points on the map  used.   If the map is not accurate
for locations of landmarks,  the  station positions determined from those

-------
                                                      Station Positioning
                                                         Method Selection
                                                              August 1986


landmarks  can  only be relocated  adequately with the same methods.  The
exact or geodetic location determined for those stations will not be correct
and other parties that try to relocate the stations using different positioning
methods may sample a different location.  Consequently,  the most accurate,
up-to-date map  of appropriate scale should be used.  Recent aerial photographs
that include the survey area can  be used to verify map accuracy.  The latest
editions  of maps  from the  U.S. Geological Survey and the National Ocean
Survey are generally of  sufficient detail and  accuracy  for positioning
purposes.   "Blue  line" maps  produced  by the U.S. Army Corps of Engineers
are made from aerial photographs  and should be used  if  available  for  the
study area because of their higher  accuracy.

     Each  station location should be plotted  and  numbered on the map.
Type of sampling, water depth, estimated time on  station, and frequency
of occupation  should be  noted.   Biological  trawl paths  and navigational
hazards should  also be indicated.  The availability  of  known targets  for
visual  or range fixes within acceptable range should be determined for
each station.  Candidate locations  for shore stations  for  each positioning
method  should  be  examined.  The selected locations should  provide coverage
of the entire sampling area.  Estimates of position errors  should be based
on anticipated LOP or angle errors  expected at  each  station.  Limiting
factors within  the survey area and  at individual  sites should be summarized,
based  on  an inspection of each  candidate shore station site.  Line-of-site
obstructions, traffic frequency,  competing transmitters, air-water boundary
irregularities, accessibility, and  security should be  evaluated.  These
evaluations are expected  to provide a basis  for judging  the feasibility
of a specified  positioning method.

     Each  station  location  is actually  an area from which a sample could
have been collected.  The size  of this area is determined by the error
inherent  in the positioning method (i.e., radial error),  the vessel offset
(i.e., distance from the hydrowire  to the location on  the vessel from which
the  station position was  determined), and the wire angle (Figure  2).  If
one assumes a zero wire angle on  the sample gear,  the  positioning  method
error  will define the area  from which the sample could  have been collected.
Any slight wire angle will  increase the size of  the  area as a  function
of depth.  The probable sampling  area  can be calculated for each specified
level of probability, depth, wire angle,  and accuracy,  as explained in
Appendix  B, "Position Error Analysis."  Probable sampling area should be
calculated for  the closest spaced  and deepest  stations at the 95  percent
probability level [the  sum of  Equation 3 (p=0.095)  and Equation  7 (wire
angle effects) in Appendix B].  This  sample area can then  be  used to determine
the minimum spacing between  stations that will guarantee statistical separation
of stations.  For studies where station overlap is not a problem, the probable
sampling  area  can be calculated  at  a lower probability (using Equation
2 [p=0.68] in place of Equation 3). For example, assuming  station  overlap
is not  a  problem (using  the sum  of  Equation 2 and  Equation 7), a vessel
using a positioning system with  an absolute accuracy of +2 m  (+6.6 ft)
in 20  m (66 ft) of water could collect  a bottom sample  from anywhere within

                                    8

-------
        VESSEL
        OFFSET
   WIRE
  ANGLE
                                                        RADIAL ERROR (RADIUS
                                                        OF PROBABLE LOCATION)
RADIUS OF PROBABLE SAMPLE COLLECTION AREA
       Figure 2.
Components  of error  in  probable area  of sample
collection.

-------
                                                       Station Positioning
                                                         Method Selection
                                                              August 1986


a 66 m2  (710  ft2) probable  sampling area.  In  100  m (3Z8 ft) of water,
the probable sampling area would increase to 419 m2 (4,508 ft2).

     Study objectives  may require a  small  probable  sampling area even if
station separation is not  a critical design  feature.  For example, a description
of seasonal variation in  benthic populations from a highly heterogeneous
area of Puget  Sound would require small  probable sampling areas.  The  size
of the probable  sampling area should be  determined by the scientific investi-
gators on a case-by-case basis.   Where  station location and depth  are  fixed
and the highest  level of probability is  assumed for the calculations, method
accuracy is the  only remaining variable.

     Generally,  it  is  sufficient to  calculate probable sampling areas at
three levels  of accuracy:   +2, 20,  and 100  m  (+6.6, 66, and 328  ft)  to
determine the  accuracy required for the  survey.  Both absolute and repeatable
accuracies of  positioning  methods can be divided into these groupings,
as will  be demonstrated in the last  section of this document, "Recommended
Positioning Accuracies."  Each positioning method  will provide accuracies
that could fall anywhere within a certain range depending on site-specific
conditions.  The +2, 20, and 100 m (+6.6, 66,  and  328  ft) accuracy  levels
are generally representative  of the highest accuracies achievable under
ideal  conditions within  the ranges  of the various positioning methods.
Candidate positioning methods can  be evaluated by  accuracy limitations
to identify the  most appropriate method.  However, state agency or  con-
tractually required accuracies should  never be exceeded.  Having established
accuracy requirements  and  survey area characteristics, the planner can
then proceed with a detailed review of  available systems.

REVIEW AND SELECTION OF POSITIONING METHOD

     Positioning methods and  associated equipment are described in detail
in Appendix C.   Additional information  can be  found in  Ingham (1975), Umbach
(1976),  Maloney (1978), Bowditch (1984), and Tetra Tech (1986).  Since
any of these methods may be appropriate for a specific  study, all  should
be considered initially.   The investigator can determine  if a  particular
method will meet calculated accuracy requirements at  the survey location
in question.   Methods that do not meet  accuracy requirements  or are inappro-
priate can be  eliminated from consideration.   The  remaining methods  should
be examined to  identify the most appropriate method for the  specific study.
A detailed evaluation of various positioning methods is presented in Appendix D.

     Each evaluation  procedure  must  be  carefully tailored to  the user's
needs and to the specific sampling or  monitoring objectives.  Factors  such
as accuracy,  range, cost (procurement or rental), portability, flexibility
(i.e., range of  conditions under which the system  can operate),  security,
user  recommendations, training, and  service  availability are basic to such
an evaluation, but are  not  all-inclusive.  The  screening of positioning
methods  in Appendix D  should be supplemented  by project-specific  considera-
tions.  Other  important  factors to consider  are  visibility restrictions

                                   10

-------
                                                        Station Positioning
                                                          Method Selection
                                                               August 1986


if night  or foul  weather work is expected, existing  equipment on the vessel
to be used, and the rental availability.  If purchase of a  system is warranted,
additional  factors  include compatibility with existing equipment, ability
to accommodate  future system expansion, and availability of ancillary  items
(e.g.,  data logger,  plotter, tracking monitor, or computer).  Potential
use of the system for  other types of projects may also be relevant.  However,
with the  expected rapid  progress  in satellite navigation systems, it may
be desirable to procure a less costly, more  labor-intensive system now
and wait  2-5 yr for  a  satellite system.  At a minimum,  the  positioning
method adopted should provide the accuracy to meet funding agency requirements
and to maintain station  separation.

IMPLEMENTATION  OF  POSITIONING METHOD

     Once a positioning method  that is adequate for the specific sampling
objective has  been selected, the proper setup, calibration, and operational
procedures must be followed to achieve projected accuracies.  If the appropriate
equipment is already  on  board the vessel or the  positioning task  is  hired
out, the  responsible party of the cruise should be sure that at least one
member of the  field crew is familiar with the positioning method.   If the
scientific team is supplying the equipment, appropriate training or experienced
personnel should be provided to ensure proper equipment operation and documen-
tation  of positioning  data.   A backup method should be available on short
notice to avoid loss  of  ship time if the primary method fails.  To  ensure
that station  locations are accurately occupied regardless of method and
that adequate  documentation is available for other parties,  record-keeping
requirements should be established, as described in the  following section.
                                    11

-------
                                                      Station Positioning
                                                            Recordkeeping
                                                              August 1986
                       RECORDKEEPING REQUIREMENTS
FIELD RECORDS
     Adequate information to ensure consistent positioning and to allow
reoccupation of stations  for  replicate sample collection or time-series
monitoring  should be kept in a  field logbook.   More detailed descriptions
facilitate problem-solving and return to the station  by the original  crew
as well as  other parties.   Required  entries into the logbook for initial
survey description, day log entries, and station  log entries are addressed
below.

Initial  Survey Description

     The positioning method and equipment  used should be detailed in the
field  logbook.  All  changes or modifications to standard methods and equipment
should  be noted  .  Names of persons  who set up and operate the equipment
should be  listed  in case questions arise.  Locations of on-board equipment
and the reference  point on the vessel where the position is actually estimated
(e.g., antennae,  sighting position) should be noted.  Distance  from the
vessel  reference point  to the  hydrowire should  be estimated to quantify
the vessel offset  component  of  positioning error.  The type of  map  that
was used  for  positioning and its  identification number or scale should
also be recorded.

     Shore station  locations should be  provided in  detail sufficient to
allow another  party to occupy the same shore station.   Shore stations should
be located  on a  monumented  point, a surveyed structure, or tied into the
local  horizontal  control at two established benchmarks by surveys performed
at the  Third-Order Class  I  level  or higher (Kissam 1981;  Umbach 1976).
Existing monumented points can be  located from National Ocean Survey Triangu-
lation  Diagrams.  Shore station descriptions should include the monumented
point reference number, location  relative to local surroundings, and  access
restrictions.   A complete copy of the surveying notes should be included
in the field logbook.

Day Log Entries

     Day log entries should record all variations from methods, equipment,
or personnel detailed  in the initial survey description.  All  problems
or irregularities should also  be  recorded along with the location,  time,
and operator.  Any weather or physical conditions that might affect achievable
accuracy  should  be  noted.   These  records help  to  explain  problems that
may not be apparent in  the field and can provide information with which
                                   12

-------
                                                       Station Positioning
                                                            Recordkeeping
                                                              August 1986
to recalculate erroneous station positions.  Relocation of  any  shore station
should be documented in the same detail  as for the initial  survey description.

     All  calibration data should be  recorded in the day log.  All manufacturer-
recommended  calibration procedures and frequencies should  be  observed.
In addition:

     •    Electronic distance measuring systems (e.g.,  microwaves)
          should be calibrated over  water between two monumented points
          at  the start and  end of  a survey.  The  calibration should
          be  performed at the same range  expected for most  stations.
          Calibration should  be checked  each day  by  recording the
          readout  at  the same  location  (single  station  systems) or
          while crossing the baseline between two shore stations  (multiple
          station  systems).

     0    Calibration of Variable Range Radar (VRR)  and Loran-C should
          be  checked  each day  by recording the  readout  at  the  same
          location.  Two VRR  ranges from the same two targets should
          be  recorded.

     t    The optical components of shore-based stations should be
          calibrated  at the start  and end  of a  survey by  recording
          the angle between the two  baselines that run from the station
          location to two other monumented  points.   This  orientation
          check  should be repeated whenever the  station  is bumped
          or  moved.

Station Log Entries

     Each station  location should be recorded in the coordinates or readings
of the method used for positioning in sufficient  detail to allow parties
not present at the survey to reoccupy the station in  the future.  The position-
ing information should be recorded at the time of sample collection (vs.  equip-
ment  deployment) and for  every reoccupation of the station, even  during
consecutive replicate  sampling.  Trawl station  positioning data  should
be  recorded  at  the beginning and end of every trawl  and at each significant
change of direction.  Supplemental positioning information  that would  define
the  station  location or  help subsequent relocation (e.g.,  anchored, tied
to northwest corner  of  pier, buoy)  should  be recorded.  If photographs
are  used  for a  posteriori plotting of stations,  the roll  and  frame numbers
should be recorded.   Depth,  time (tidal height), ship heading, and wire
angle estimation  should also be recorded  for each  occupation of a  station.

     Conditions  at  each station should  be recorded  and checked against
the range of  operating  conditions known to  produce optimal  accuracy for
the specified method.  Examples of these  specific  requirements  are presented
below for common  positioning methods:
                                    13

-------
                                                       Station Positioning
                                                             Recordkeeping
                                                               August  1986
     t    VRR - Three  fixes  from permanent objects should be recorded
          for each station.  Fixes should all be  from  targets 0.16 to
          6.4 km (0.1  to  4 mi)  away and all from the same range  scale
          on  the VRR.   Station positions should  be  supplemented with
          Loran-C coodinates, depth, and all visual line-of-site  fixes
          available.

     •    Loran-C - The range of Loran-C coordinates for the  survey
          area should be recorded.   This  helps the  operator notice
          signal  interference  problems  and signal  jumps (10-skip)
          that may otherwise result  in recording  faulty coordinates
          for  a  station position.   Both  day and  night coordinates
          should be recorded from one location inside the study area
          to  check for  a 0.1 usec day/night variation.  Any fluctuations
          at  the 0.1 usec level while on station should  be recorded,
          including the reading that  appeared on  the display the majority
          of  the time.

     •    Microwave systems - Crossing angles of  LOPs should be between
          35  and 150 degrees  for all  stations.  Readout should be
          noted two to three  times  before being recorded to eliminate
          multipath errors.

REPORTING REQUIREMENTS

     Sampling data reports supplied to other parties should include rationale
for selecting the positioning method  and recordkeeping standards observed
during data collection.  Any specific problems (e.g., wind, currents,  waves,
visibility, electronic  interferences) that resulted  in positioning problems
and those stations affected should be identified.  Estimates of the accuracy
achieved during the survey should be  included.  This information  will help
the reader to determine the quality and consistency of the sampling.

     Station  locations  should be reported in State Plane coordinates (specify
north or south zone) or by latitude and longitude (to the  nearest second).
Coordinates  need not  be  reported  for each  replicate collected; a  single
set of coordinates for  the station is sufficient.  Depth  corrected to mean
lower  low water (MLLW) should  be  supplied for each station.  Coordinates
for the start and end of each trawled section should be provided.
                                   14

-------
                                                       Station Positioning
                                                      Recommended Accuracy
                                                               August 1986
                    RECOMMENDED  POSITIONING ACCURACIES
     Accuracy  requirements  for  different types of environmental sampling
requirements are  examined in this section.   Sampling that  requires  highly
accurate  positioning methods to help  ensure quality  and  consistency in
collected  samples and  in data interpretation and  analysis is described.  The
corresponding acceptable positioning methods  are identified.  This process
is presented for  two specific sampling objectives identified by  the  Puget
Sound Dredged Disposal Analysis (PSDDA).

CLASSIFICATION OF SAMPLING

     The sampled  medium and collection method can be used to classify sampling
activities.   The  collected sample will be representative  of either a  point
location or  a larger area.  Examples of point and areal samples are provided
in Table 2.   For  purposes of identifying  accuracy requirements for sampling
activities,  samples collected from a homogeneous  medium (e.g., water column)
are considered representative of a larger area  and thus similar to an  areal
sample.   A  third classification of sampling activities which could include
either point or areal  samples is  also considered in Table 2.  This  third
group  (referred  to in this document as special  studies) consists of studies
usually used  to identify induced effects on the environment.  Characterization
of natural  variability will  be important for special studies and point
samples collected from heterogeneous environments.   Accuracy of the positioning
method  (i.e.,  error associated with station location, or spatial  resolution
of the location from which the sample could have  been collected) will affect
the ability to  identify  the difference  between natural  variability and
induced effects.   Other sampling will not require this  level  of  accuracy.
Therefore,  levels of positioning accuracy can be recommended for the three
classifications of  sampling that are identified from Table 2  (special studies,
point  sampling at the sediment boundary, and areal sampling including any
samples from the  water column).

RECOMMENDED  ACCURACY

     Accuracy  levels of the methods detailed in  Appendix C and evaluated in
Appendix D can be categorized as  +2,  20, or 100 m (+6.6,  66, and  328 ft)
for the  reasons  discussed  earlier in this document.  The highest accuracy
level  [+2  m  (+6.6 ft)] is recommended for sampling performed under  special
studies  in  Table 2, point sampling studies in  Table 2, and studies designed
to build  the developing  database  for  the Puget Sound Estuary Program.
The highly accurate station locations possible  using +2-m (+6.6-ft) accuracy
will  allow more closely spaced stations and better detection of  trends or
gradients that could otherwise be masked by  variance in  replicate samples
collected  using less accurate positioning methods.  In addition, uniformity


                                    15

-------
           TABLE 2.  CLASSIFICATION OF COMMON TYPES OF SAMPLING


Point Sampling

     Water Column3

                    Current measurements
                    Water samples
                    Sediment traps

     Sediment Boundary

                    Sediment physical/chemical
                    Bioassay sediments
                    Interstitial  waters
                    Benthic infauna
                    Bivalve coring

Areal Sampling

     Water Column

                    Plankton tows
                    Pelagic fish trawls

     Sediment Boundary

                    Demersal fish trawls
                    Crab pots or trawls
                    Bivalve dredging
Special Studies'*
     Point source investigations
     Time series sampling
     Synoptic collections
     Spatial gradients (including surface microlayer,  dye studies)
     Station grids
     Dredging site sediment characterizations


a Homogeneous nature of water column makes all  water column sampling  equivalent
to areal sampling.

b Studies in this classification may  contain any  of the  sampling types
listed above but are considered separately because  of additional  accuracy
requirements.
                                  16

-------
                                                      Station Positioning
                                                     Recommended Accuracy
                                                              August  1986


of positioning accuracy across  these studies is  expected to improve compara-
bility of  data and is one of the  factors that affects overall  data quality.

     However, if the  location or variable  sampled  does not require the
highest accuracy levels, a  lower accuracy level [+20 m  (+66 ft)] may be
adequate.   For  example, results  from  sampling the relatively homogeneous
benthos of deeper waters would  probably not be adversely affected if  the
replicate  grabs were collected  from a larger area.   If +20-m (+66-ft) accuracy
is sufficient to address the objective in question,  and the funding agency
cannot  justify  additional  funds  for  the use of  higher accuracy in order
to conform  to the  database requirement, the lower  accuracy may be appropriate.
In many cases,  point sampling  studies could be  adequately performed at
the lower  level  of accuracy.  Such  exceptions should be evaluated on a
case-by-case basis.

     The +20 m  (+66  ft)  accuracy  level should  be adequate to collect areal
samples.  The more homogenous nature of the medium and  larger area sampled
make higher  accuracies unnecessary.  The exception would be water samples
collected  near point sources where location is important in determining
dilution  characteristics.  The lowest accuracy  level  [+100 m (+328  ft)]
generally  is not recommended.

RECOMMENDED  POSITIONING METHODS

     Positioning methods  that can achieve  the two  recommended  accuracy
levels are presented in Table 3.  Approximate ranges of both absolute  and
repeatable  accuracies for  some of  the methods are compared in Figure 3.
Methods with +2  to 10 m  (+6.6  to 33  ft) accuracy  are acceptable for  the
higher  level of accuracy.  Those  with +20 to 40 m  (+66 to 132 ft) accuracy
are acceptable for the lower level  of accuracy.  In addition, some of  the
advantages  and  disadvantages for these positioning methods are presented
in Table 3.  Absent from Table  3  are some of the optical positioning methods
discussed in Appendix C.  While  these methods do not  offer the accuracy
necessary  for most sampling farther from shore,  they can be highly accurate
in areas  within 0.5 km (1,640 ft) of shore.  Along the waterfront of urban
areas and  in waterways where other methods may be unsuitable because of
interference or minimum range requirements, optical positioning techniques
may be preferable.  Optical range finders and visual  line-of-sites  can
provide accurate positioning if  detailed notes are  taken.  Positions may
also be determined by taking three photographs  to  allow  for a posteriori
triangulation from aerial  maps or  "blue lines."  Any positioning by optical
methods is improved if multiple methods  (e.g., photographs) are employed.

     The  lead scientific investigator  should decide whether the limiting
factor in  study  design is  absolute or repeatable  accuracy, what effects
the  location will  have on  positioning capability, and determine which one
of the candidate methods is appropriate  for the specific  sampling program.
The  following  determinations of  appropriate accuracy levels for specific
                                    17

-------
                                           TABLE 3.   SUMMARY OF ACCEPTABLE  POSITIONING  METHODS
Category
Higher Accuracy
Group
Theodolite








Representative Achievable
Equipment Accuracy


Table C-2 10-30 sec
Table C-3 * 1 m and up







Maximum
Range Cost Advantages


<5 km $1,000-$4,000 Traditional
method.
Inexpensive.
High accuracy.
Successfully
applied.
Restricted areas.


Disadvantages


Line-of-sight.
Two manned shore
stations.
Simultaneous
measurements.
Limits on inter-
section angles.
Area coverage;
station movement.
            EON I
Table C-7
1.5-3.0 cm
 3 km without  $3.500-$15,000
multiple prisms
oo
            Total Stations
Table C-8
  5-7 cm
    <5 km     $8,000-$30,000
            Microwave
            Navigation
            Systems
            Range-Azimuth
            Systems
            Satellite Systems
Table C-9
                                                      1-3 m
                     25 km     $40,000-$90,000
Table C-12     0.01° and 0.5 m
                                  Table C-ll
                   1-10 m
                     <5 km    $65,000-$100,000
                   (optical)
                     30 km
                     (elec)

                      none     $150,000-$300,000
                               (initial  units)
                             under $10.000 later
Extremely accur-
ate.  Usable for
other surveying
projects.  Cost.
Compact, port-
able, rugged.
Single onshore
station.  Other
uses.  Minimum
logistics.
                                No visiblity
                                restrictions.
                                Multiple  users.
                                Highly accurate.
                                Radio line-of-
                                sight.

                                High accuracy.
                                Single station.
                                Circular  coverage.
                                High accuracy.
                                Minimum logis-
                                tics.  Use in
                                restricted/con-
                                gested areas.
                                Future cost.  No
                                shore stations.
Motion and
directionality of
reflectors.  Line-of-
sight.  Visibility,
unless microwave.
Two shore stations.
Ground wave reflec-
tion.

Reflector movement
and directionality.
Prism costs.  Line-
of-sight.  Optical
or infrared range
limitations.

Cost.  Multiple
onshore stations.
Logistics.  Security.
Signal reflective
nu11s.
                   Single user.  Cost.
                   Line-of-sight.
                   Signal reflective
                   nulls.

                   Current coverage.
                   Initial develop-
                   ment cost.

-------
TABLE  3.   (Continued)
Category
Lower Accuracy
Group
Sextant3
Representative
Equipment
Table C-4
Achievable
Accuracy
+ 10 sec
+ 2 m and up
Maximum
Range
<5 km
Cost Advantages
$1 ,000-$2,000 Rapid. Easy to
implement. Corn-
Disadvantages
Simultaneous
measurement of two
Loran-C
Table C-ll
15 m and
  up
                                                            none
$1.000-J4,000
Variable Range
Radar
                       Table C-5
                    0.5°
                    25 in
                                                           16-72 km
                               (4.000-$10,000
                                                                                        iron equipment.
                                                                                        Low cost.  No
                                                                                        shore party.
                                                                                        High accuracy
                                                                                        closer to shore.
No visibility or
range restric-
tions.  No ad-
ditional person-
nel.  Low cost,
existing equip-
ment.
                No visibility
                restrictions.
                No additional
                personnel.  Low
                cost, existing
                equipment.
 a  Accuracies  greater than  +20 m are not  common farther than 1 km from shore  under
 normal operating conditions and is therefore considered in the lower accuracy group.
angles.  Target
visibiIities,
location, mainte-
nance.  Line-of-
sight.  Best in
calm conditions.
Limits on accep-
table angles.

Interference in
some areas.  Used
only for reposi-
tioning except  in
Iimited areas,
need to locate
station initially
with another
system.

Line-of-sight.
Re Iies on map
accuracies of
targets.  Ac-
curacy decreases
with range scale.

-------
ro
o
SEXTANT
     OPEN WATERS
     NEAR SHORE
VARIABLE RANGE RADAR
LORAN-C MAPPED AREAS
VRR, VISUALS, LORAN-C
     OPEN WATERS
     NEAR SHORE
MICROWAVE SYSTEMS
SATELLITE SYSTEMS
RANGE-ANGLE SYSTEMS2
REPEATABLE
ACCURACY
                                               20        40         60        80       100
                                                  APPROXIMATE ACCURACY RANGE (±m)

                                       a  INCLUDES TOTAL STATIONS AND RANGE-AZIMUTH SYSTEMS
                                                                                 120
                   Figure 3.  Comparison of  approximate accuracy ranges of some positioning methods.

-------
                                                      Station Positioning
                                                     Recommended Accuracy
                                                              August  1986


sampling  objectives are presented  as  examples of the process.  Sampling
studies conducted for PSDDA will  consist of two basic objectives:

     •    To characterize the sediments to be dredged

     •    To monitor disposal-related  impacts and trends at dredged
          material disposal  sites.

Both objectives are included under special studies.  The higher accuracy
is recommended  for sediment  characterization because exact  locations of
the areas  to be  dredged must be defined for later removal of only the specified
sediments. Dredged areas are usually close to shore.  Most of the higher
accuracy methods in Table 3 will  meet these requirements close to land.

     High  repeatable  accuracy is required  for site  monitoring to ensure
that observed trends are  temporal  and not spatial.  Most of these sites
will be over 2 km from  shore.  Range  restrictions may  preclude some of
the higher repeatable accuracy  systems (see Appendix  D; Figure D-l) at
selected  sites.  Absolute position  accuracy may not be important, although
this determination will  have to be  made for each disposal  site and for
individual resources of concern.   If  absolute positioning is needed the
higher accuracy  methods are recommended.
                                   21

-------
                                                       Station Positioning
                                                               References
                                                              August 1986
                               REFERENCES


Bowditch, N.   1984.  American practical  navigator.  An epitome of navigation.
Defense Mapping Agency Hydrographic/Topographic Center, Washington,  DC.

Holme, N.A., and A.D.  Mclntyre (eds).   1984.   Methods  for the study of
marine  benthos.   Blackwell  Scientific Publications.  Oxford, U.K.  387
pp.

Ingham,  A.E.  1975.  Sea  surveying.  John Wiley  and Sons, New York, NY.
306 pp.

Kissom,  P.   1981.  Surveying  for civil engineers.  McGraw-Hill Book  Co.,
New York, NY.

Maloney,  E.S.  1978.  Duttons navigation and  piloting.  Naval Institute
Press, Annapolis,  MD.  910  pp.

Tetra Tech.  1986.  Evaluation of coastal  survey positioning methods for
nearshore marine and estuarine waters.  U.S. EPA  Office of Marine and Estuarine
Protection, Washington, DC.  Tetra Tech, Inc., Bellevue, WA   107 pp.

Umbach, M.J.  1976.  Hydrographic manual, fourth  edition.   U.S. Department
of Commerce, NOAA, Rockville, MD.  414 pp.
                                    22

-------
             APPENDIX A






SITE-RELATED POSITIONING LIMITATIONS

-------
                                                       Station  Positioning
                                     Appendix A - Site-Related  Limitations
                                                               August 1986
                               APPENDIX A

                   SITE-RELATED POSITIONING LIMITATIONS


     Location has many effects on the  accuracy and applicability of various
positioning  methods.  Such effects are  described  below for optical,  radar
range,  and  short-range  and  long-range  electronic positioning  methods.
These methods  are described because they are commonly used in  Puget Sound.

     Optical  positioning methods  rely on the visual resolution of objects
with a known position.  Built structures provide  more accurate  fixes than
land  features because sharply defined  objects provide better  resolution.
The ability to resolve an object decreases  with distance.  Within 5 km
(3.1  mi)  of the shoreline and in more developed areas, accuracy of optical
methods can  be comparable to that of  electronic methods (Umbach  1976).
However, optical methods are more dependent upon proper operation and  target
choice than  are other methods.  Urban embayments and waterways are  therefore
better  suited for optical  positioning  than  are regions in central Puget
Sound or areas along less populated, featureless shorelines.   The  abundance
of accurately located channel markers throughout Puget Sound provides good
sightings in otherwise  featureless areas.  On featureless  shorelines, a
line tangent to  the shoreline is used as a line-of-position, with considerable
reduction in accuracy.  Use of optical   methods is restricted to  daylight
hours of good  visibility.

     Positioning  by  multiple  ranges  measured from a variable  range radar
system requires  fixes on known positions, but eliminates visibility restric-
tions.   However,  because  the radar signal  is  shadowed beyond  the first
object it strikes,  the choice of targets can be  limited.  A second  limitation
is possible misidentification of reflection sources in developed  areas.
Positions based on  misidentified reflection sources are inaccurately located,
but can be reoccupied if the same perceived reflection source  is  subsequently
used.  A third target  should be used  to crosscheck position  determined
from  two other targets.  All three fixes should  be on the same  radar range
scale.

     Reflections depend on target position and  alignment.  The most accurate
radar range fixes are based on reflections from objects 0.16-6.4 km (0.1-4  mi)
distant (Crawford, P.,  personal communication).  At stations farther  from
shore, adequate targets  may not  be available.  Reflections closer than
0.16  km (0.1  mi)  may be erroneous and  should  not  be  used.  Sloped  headlands
and tidal flats  are not usable as targets.

     Short-range  electronic  positioning systems  (e.g., microwave)  involve
at least two shore stations (transponders)  and an  on-board transmitter.

                                   A-l

-------
                                                       Station  Positioning
                                     Appendix A - Site-Related  Limitations
                                                               August 1986


A set of  stations is required  for each study area.  Successful  reception
of electromagnetic signals is the critical feature for effective operation
of electronic positioning systems.   Signals from the transponders or shore
stations should  be received at an angle of  30  to 150 degrees;  90 degrees
is optimal.   Signal  reception is dependent upon electronic "line-of-sight"
and may be blocked by tight quarters (e.g., waterways, rivers, and  shorelines)
and heavy vessel traffic.   Such problems are  alleviated by  newer, more
expensive systems that accommodate over  a dozen transponders.  With proper
transponder  locations,  microwave systems  can be used to position a vessel
any distance  from shore in Puget Sound  except  closer than 100  m (328 ft)
to one of the transponder locations.

     Accuracy of microwave  systems  depends upon placement  of the remote
transponders  or  shore stations.  For example,  remote stations not located
on a  monumented  point will  increase  error  of  the vessel position  fix.
However, access restrictions, benchmark locations, and line-of-site considera-
tions  limit  the available  transponder locations and achievable angles.
Permanent shore  station locations are further  limited by  availability of
power sources and site security.

     Repeatable  accuracy  of microwave systems, while not affected by transponder
location errors, is dependent upon  the 1ine-of-position  angle.  Certain
combinations  of  transponder and vessel locations may result in  signal cancel-
lation (range holes)  and failure to obtain  a fix.  Occurrence of range
holes varies  by  location, is  impossible  to  predict, and may force relocation
of a shore station.   In  developed areas,  reflections from metal objects
or buildings  may compound the problem or cause  jumps in the received signal.

     Long-range  electronic positioning systems operate on permanent transmitting
stations and  user-carried receivers.  The only  receivable long-range system
in Puget Sound  is Loran-C.   However, because  land masses distort signal
propagation,  Loran-C  charts are of unknown  accuracy  in inland waters such
as Puget Sound.  In addition, unidentified  electronic  sources interfere
with Loran-C reception  in some areas and prevents  its use  much of the time
(Figure A-l).   However, Loran-C is accurate  for  repositioning at locations
where readings were  recorded  in Loran-C on original occupation.  Some areas
may exhibit a day/night signal variation of about  0.1 usec  (Eaton C., personal
communication) and should be  checked when sampling at night.

     Recent  Loran-C maps based on comparison  of Loran-C coordinates  with
those from other methods at the  same  stations  support  positioning to  a
resolution of  0.1 usec  of  the Loran-C  signal   [about  37 m  (120 ft)]  for
limited  areas in the  vicinity  of  Elliott Bay (Sturgill, D.,  personal communi-
cation).  Accurate  maps for other  areas  are not available.   Distortion
can be defined  for areas outside  Elliott Bay only  by  taking  readings  at
benchmarks around  the shoreline.   Accuracy is  less reliable with distance
 from the  benchmarks.   Therefore,  while Loran-C  is  usable around Elliott
Bay for  initial  positioning,  it  can be  used in other  areas  only  as a reposi-
tioning  tool  and will require defining  Loran-C station  coordinates.


                                    A-2

-------
LARGE AREA REGULARLY AFFECTED
 BY POSSIBLY A SINGLE POWERFUL
                      SOURCE
  OCCASIONAL SHORT-TERM INTER-
        FERENCE IN ELLIOTT BAY
   SEOLA BEACH TO PT. WILLIAMS
      POSSIBLE AIRPORT SOURCE
    10
NAUTICAL MILES
      KILOMETERS
    20
«
                                                  NOTE: MOST OF THE SOUTH SOUND
                                                  ALSO EXHIBITS INTERFERENCE FROM
                                                  A SINGLE SOURCE NEAR TACOMA
  Figure  A-l.  Regions of LORAN-C  signal  interference in  Puget
                Sound.
                                  A-3

-------
                                                       Station  Positioning
                                      Appendix A - Site-Related Imitations
                                                               August 1986
                               APPENDIX A

                               REFERENCES
Crawford,  P.  15 November 1985.  Personal  Communication  (phone by Jeff
H. Stern)  University of  Washington Research  Vessel  Clifford  A.  Barnes.
Seattle, WA.

Eaton, C.   9 December 1985.  Personal  Communication (phone by Jeff H. Stern).
Kittiwake  Research Vessel.  Seattle,  WA.

Sturgill,  D.  18 November  1985.   Personal Communication  (phone  by Jeff
H. Stern).   Municipality of Metropolitan  Seattle, Seattle, WA.

Umbach, M.J.  1976.  Hydrographic manual,  fourth edition.  U.S.   Department
of Commerce, NOAA, Rockville, MD.  414 pp.
                                    A-4

-------
      APPENDIX B





POSITION ERROR ANALYSIS

-------
                                                       Station Positioning
                                               Appendix B - Error Analysis
                                                              August 1986
                               APPENDIX B

                         POSITION ERROR ANALYSIS


     Errors have been traditionally classified in three categories:  gross,
systematic,  and  random (Mikhail  and Gracie 1981).  Gross errors  are mistakes
due to observer  carelessness and must be detected and eliminated from positional
measurements.  Systematic errors occur due  to  a  deterministic  system, in
this case  the sextant,  range  finder, or electronic positioning equipment.
Such errors  can  be expressed by functional relationships and can  therefore
be corrected for in the final measurement.   Random errors reflect variations
in measurements  that  remain after eliminating gross errors and  allowing
for  systematic  errors.  As  their name implies, such observational errors
are random  and have no known functional  relationship.  Random  errors, the
subject  of  the  following discussion, are dealt with on a statistical  basis
using probability models.

     All positional  fixes are  in error  to some extent.  This  is  true even
if the transmitting stations have been very accurately  located  and bias
or systematic errors in position coordinates  have been eliminated.  Random
errors will  still cause  a variable  displacement of position  lines about
their  computed  positions.   Each method used to  establish a  position has
inherent random errors  that determine the overall accuracy of a fix.  Absolute
or predictable  accuracy refers to  a method's ability to correctly define
a position by latitude and longitude (Bowditch 1984).  Repeatable or relative
accuracy measures a method's ability to return  the user to the same position
time after time.  The  difference between these two accuracies can be signifi-
cant.   Therefore, it is important to identify which accuracy is of  concern
for positioning  during a  specific sampling program before determining the
error associated with that  accuracy.

     Positional accuracy  is usually stated  in terms of the probability
of being within a certain distance of  a desired geographical  location.
The  line-of-position (LOP) can therefore be considered a strip whose  width
is a function of the  LOP  uncertainty, which  can be described by its standard
deviation  (o)  (Figure  B-la).   Because  a  fix involves  at least two  LOPs,
each having an uncertainty,  the LOP crossing boundary forms a parallelogram
(Figure B-lb).   The parallelogram  area  is dependent on  the size of the
standard deviations  of  the two  LOPs and the angle at  which  they cross.
The  parallelogram  area is given by:

               A =  (2oi)(2o2) esc* =  4oio2 esc*                         (D
where 01 and 02 are  the LOP measurement standard  deviations.  Because the
cosecant function becomes large  for  angles less than  30 degrees and  more
                                    B-l

-------
                                  .MEASURED LINE OF
                                  'POSITION NO 2
                                   MEASUREMENT
                                   VALUE SPREADS
                                                  MEASURED LINE OF
                                                  POSITION No 1
   SHORE STA. 1
BASELINE
                                                         SHORE STA. 2
a.
         ERROR
  PARALLELOGRAM
b.
                                      CIRCULAR PROBABLE
                                      ERROR
                                           SOURCE: MODIFIED FROM BOUDITCH 1984
     Figure B-l.  Line of  position measurements for two shore
                   stacions  depicting LOP uncertainly (a),  and
                   associated  error indicator  (b).
                                  B-2

-------
                                                      Station Positioning
                                              Appendix B - Error Analysis
                                                              August  1986


than 150  degrees, so too does  the size of the error parallelogram (Figure B-2).
Thus,  to  reduce positional  uncertainties,  shore station  locations should
be located  such that very large or very small cut angles do not occur within
the survey  area.

     Probability theory indicates a 68.27  percent chance of being within
+1.0o of  the mean of a set of  normally distributed measurements.  However,
the probability  that the measured  position is within  the  parallelogram
is less.  For two sets of measurements,  the joint or  product probability
is  (0.6827)2  x  100 or 46.6  percent.   A  probable parallelogram is defined
as one for  which the probability of a measurement being within its boundary
is  50 percent.   This occurs  at distances slightly larger than +1.0o  (i.e.,
+1.05o).  Using +2.0o results  in  a  parallelogram within  which it is 91
percent probable that the positional location occurs.  The principal drawback
is that the error parallelogram  is  not a curve of equal  probability.  An
equal probability contour from a two-LOP fix is an ellipse centered on
the LOP intersections.  As with the parallelogram, the area and configuration
of the error ellipse are dependent on the crossing angle  and standard deviations
in each set  of measurements.  The ellipse in which 50 percent of the expected
measurements  lie is referred to  as  the  probable ellipse.  Techniques for
determining the orientation  and  the magnitudes of an  error ellipse  for
various probability levels are presented by Mikhail and Gracie (1981).

     It  is more common  to establish the  probability that a  position is
located within a circle of  a particular radius than  to define an error
ellipse.   The  resultant circular probable error  (CPE or CEP) is the  radius
of a circle  within which 50 percent of the fixes should  be  located.  Bowditch
(1984) provides  two methods for  determining the sizes  of  error circles
at a desired probability level,  given the standard deviation in each LOP
measurement  and  the associated  pattern crossing angle.  The probability
that the  measured position  is  within a circle of a  selected radius can
also be determined.

     When  the  standard  deviations of  two sets  of  measurements are equal
and the cut angle  is 90 degrees,  the error figure becomes  a circular normal
distribution.   Table B-l can  be  used  to determine  the  probability that
location  is within a circle  of  given radius  from the measured  position.
At  a  radial  distance (R) from  the  true position equal  to lo (i.e., the
ratio of  R/o=l), there is  a  39.9  percent chance of being located within
the circle (Table B-l).  A 50  percent  probability occurs for a circle with
a l.Zo radius  (R/o=1.2), and a 90  percent probability  occurs for  a circle
with  a 2.2a radius.  These  probabilities change as  the cut angle varies
from 90 degrees.   For example,  if the cut  angle is 40 degrees,  a radius
greater  than  3.4o is needed to  achieve a 90 percent probability that the
position  lies  within the circle  boundary.   Table B-l  can also be used  to
determine if positioning error is  small  enough to separate the  error circles
of  two closely spaced positions.   The  probability of  being located within
a circle can  be calculated  (at  any cut angle) for an error  circle radius
                                   B-3

-------
                                                              AREA OF
                                                             UNCERTAINTY
CD
i
                              /    30°
                                ANGLE OF
                             /     CUT
                                                                            90°
                                                                         ANGLE OF
                                                                            CUT
                                                        POSITION,
                                                          LINES
                       Figure B-2.   Angle-of-cuc effects CM. fix  accuracy.

-------
             TABLE  B-l.    PROBABILITY  VERSUS  R/o  FOR  ELLIPTICAL  BIVARIATE
                      DISTRIBUTIONS  WITH  TWO  EQUAL STANDARD  DEVIATIONS
                                II"
                                      10*
                                      II"
                                                   4*'
                                                         41"
                                                         M"
                                                                      II"
                                                                            M"
                                                                                  41"
                                                                                         7*"
                                                                                                       4*"
                                                                                                              II*
11
4 0

\\
4 1
4 4
      .001
      i
      .01$
      .051
      .IS!
      .0*1
      .0**
      .07*
      .07*
.0*1

!io*
.10*
.us
.11*
.124
.12*
.IM

'.14*

:iJS
.IS*
.1*4
.1*9
.174
.179
.111
 II*
.191

 ,6*1
             .011
             .01*
             .02*
             .01*
             .OU
             .0*0

             .070
             .ON
             .0*2
             .10*
             .IIS
             .12*
             .IM
             .1*1
             .IS*
             .1*9

             .17*
             .119
             .1*9
             .209
             .21*
             .22*
             .21*
             .241
             .2U
             .2*0
             .277
             .2*7
             .2**
             10*
             .US
             .124
             .111
             .141
.8*
             .00*
             .012
             .021
             .02*
             .041
             .057
             .072
             .OU
             .10*
             .120
             .11*
             .151
             .171
             .IM
             .204
             .220
             .21*
             .251
                   .124
                   .IM
.IT*
.1*1
.406
.41*
.412
.441
.451
.470
.412
.4*4

.IS
.529
.141

•H
            .007
            .015
            .027
            .019
            .057
            .075
            .095
            .11*
            .117
            .151
            .III
            .204
            .22*
            .247
            .2*9

            !no
            .12*

            .M*
            .1*7
.422
.440
 4S7
.474
.4*0
.SOT
.522
.SV
.SSI
.5*1
.511
.5*7

;*2s
.611
.651
.6*4
.676
.6M
.700
.01*
.011
.041
.070
.0*1
.117
.141

.16*
.196
.224
.251
.271
.104
 110
.155
.MO
.401

.42*
.441
.47C

!su
.512
.SSI
.570
.SI*
.607

.624
.•41
.457
.671
                   .711
                   744
                   .7S7
                   .770
                   .701
                   .791
.011
.022
.01*
.057
.0*2
.110
.IM
.169
.199
.211
.2*4
.296
.127

.151

:!&
 445
.471

.491
.571
.54*
.570
.592
.614
.614
.654
.674
.692

.710
.727
.741
.759
.771
.711
.101
 III
.12*
.IM
.141
.»59
.169
 •71
                         .Oil
                         .025
                         .045

                         .sn
                         .19*
                         .221
                         .26$
                         .102
                         .119
                         .17*
 442
.471
.SOS
.51*

 5*1
 U9
.615
.619
 Ml
 6te
 70*
.72*
.74*
.76J
.711
.797
 • 12
.12*
.140
.•52
.1*4
.17$
.M*
 I9S
.904
.911
.012
.021
.050
.074
.IM
.141
.177
.217

.25*
.296
.117
.171
.417

.455
.491
.52*
 S59
.ill

.621
 649
 67$
.700
.724
.71*
.76*
.71*
.10$


 •IT
 •5.'
.I*'.
 171
.119
.9UO
.910
.919
.927
 915
.942
 9M
.954
.VU
                               .014
                               .011
                               .oss
                               .001
                               .117
                               .155
                               .19$


                               .211
                               .12$
                               .170
                               .414
                               .45*

                               .497
                               .51*
                               .571
                               .6(1*
                               .611

                               .672
                                7
-------
                                                      Station Positioning
                                               Appendix B - Error Analysis
                                                              August 1986
of one half  the distance between  the two positions  to determine if position
overlap may occur statistically.

     Positional accuracy is often quoted  using radial error, rms error,
or d   ,  all of which have the same meaning.  As with CEP, the error figure
is ar(?ftcle having a radius equal  to:
                              rms  =  (ox2 + oy2)                        (2)

where ox  and  oy are the lo error  components along the major and minor axes
of the probability  ellipse (Figure B-3).  This  is  often referred  to as
the Id    or lo  fix  accuracy.  Similarly, a 2o fix is  given  by:

                                [(2ox2) + (2oy2)]1/2                    (3)

Given  the LOP standard deviations oj and 02, the standard  deviations along
the basic ellipse axes can be determined by:
oK2 = (1/2 sin2$)  { o j2 + o22 + [(c^  + o22)2 - 4(sin0)  0]a2)     (4)

and

o2 = (1/2
The  probability of  being within a Id   circle varies from 63  to 68 percent
depending on the ellipticity of the^ error ellipse.   For a  2dpms circle,
the probability varies from 95 to 98 percent.

     It is common  to describe  two-dimensional error distributions by two
separate  one-dimensional  standard deviations  along each error axis (ci
and o2).  Error  circles  about a measured  position can be constructed for
different LOP  standard  deviations  and cut angles.  This  can  be  done by
first  calculating  ox and  oy (the standard deviations along  the major and
minor axes of the error ellipse) knowing the crossing angle and the standard
deviations in each measurement (o\ and o2).   Table  B-2 is then used to
provide probabilities as a  function of c,  the ratio of the smaller to larger
of ox  and cy, and  K as the  ratio of the error  circle radius  to the larger
of ox  or  cv.  For  example,  for a 50° angle of cut,  a °i = 15 m (49 ft)
and  c2  =  20 m (66  ft), the  probability of  location within a circle of 30 m
(98  ft) is:

     (Ol)2 = 225 m2
     (o2)2 = 400 m2
     sin2* = 0.5868

Substituting into equations 4 and 5, ox is 29.9 m (98.1 ft) and oy  is 13.1 m
(43.0 ft).
                                   B-6

-------
                                           SOURCE:  BOWDITCH 1984
Figure B-3.   Illustracion  of radial error.
                             B-7

-------
      - B-2.  CUCJLAR ERROR PROBABILITIES  AS  A  FUNCTION OF MEASUREMENT
            STANDARD DEVIATION RATIOS AND ERROR  CIRCLE  RADIUS

\
•X
0. 1
0.2
0 3
0 4
as
0 6
0 7
0 8
0 9
1 0
,
i
3
5
6
7
8
9
2.0
2. 1
2.3
2.3
2 4
2 5
2.6
2 7
2.8
2 9
3 0
3*3
3 3
3 4
3 5
3 6
3 7
3 8
3 9
4 0
4. 1
4 2
4 3
4. 4
4 5
4.6
4 7
4 S
4 9
j 0
5 1
5 2
i 3
5 4
3 5


J '
3 3
5 9
4 0



0 0
0796557
1585194
2358228
J 108435
3829249
4514938
5160727
5762892
6313797
58208S5
7286679
769S607
3063990
8384867
3663856
S9040U
9108691
9281394
9425669
9544997
9442712
9721931
9785518
9836049
9875807
9906776
9930661
9940897
9962684
9973002
99S0648
9986257
9990332
9993261
9995347
9998818
3997344
999&J53
9999038
9999367
9999587
9999733
9999'2*
9999*92
9999933
9999958
9999974
9*999*4
9999990
9999994
9999998
9999999
' 9999999
1 0000000
1








a i
0443987
1339783
2211804
3010228
3755E84
4457708
5115048
5725957
6288721
6802325
7266597
76X2215
3050848
.0374049
9655127
8897008
9103102
9274964
9422182
9542272
9640598
9720304
9784275
9835108
9875100
9908249
9930271
994S612
9962477
9972853
99S0542
9986182
9940279
9993225
9995323
9996801
9997332
9993545
9999033
9999363
9999585
9999732
9994828
9999891
9999931
9999957
9999974
i 99999*4
9999990
9999994
9999997
9999998
9999999
9999999
1 0000000

1







0 2
0242119
0884533
1739300
2635181
34S1790
• 25S60S
4960683
5604457
6191354
G723586
7202682
7630305
8008554
8140018
9627728
8875060
9085619
9263I2S
9411299
9533775
9434011
9715237
9780408
9832180
9872900
9904612
9929062
9947727
9961934
9972391
9980212
9985949
99901 16
9993112
9995245
9996748
9997747
9998522
9999018
9999353
9999578
9999727
9999826
9999889
9999931
9999957
9999973
9999984
9999(90
9999994
9999997
9999998
9999499
9999999
1 0000000









0.3
0164176
0128396
1318281
2139084
3003001
3846374
4633258
3349387
5993140
6568242
7079681
7532175
7929968
8277048
8577362
8834914
9053766
9237989
9391586
9518415
9622127
9706109
9773450
9820918
9868953
9901674
9926894
9946141
4940684
9971564
0979622
99D5533
9912909
9993105
9996653
•1997733
9"93478
9V9S089
9999334
9999566
4499720
9999821
9999386
9999929
9999954
9949973
1 9999983
9999990
9999994
9999996
9999998
9999999
9999990
1 0000000





i



1
0 «
0123875
0432413
1039193
1743045
2532951
3357384
4170862
4941882
3651564
6291249
6859367
7359558
7793550
9169851
8493071
3768444
9001746
9197275
9359855
9493815
9403170
9691397
9762419
9819594
9862720
9897043
9923483
9943649
9953S79
9970266
4979699
9944480
99)9368
9492513
9994088
9998505
9097633
99934 1 2
9490945
9W9305
99*9547
9999707
9494813
9999881
9999925
9999954
4999971
9999983
9999990
9999994
> 9909996
9999998
9999999
1 9999999
I 0000000





1



0 5
0099377 '
0340193 i
083153.1
1451308
21528SO
2914682
3099305
4474207
3211998
3900953
6524489
7079973
7567265
7989288
8350816
8657559
8913536
9130680
9308615
9454546
9573205
01.60345
4745239
9805703
9853112
9889934
991*260
9439842
9936126
9468294
9977298
9403392
40..H677
9992113
9994559
9996281
9997432
9990111
9999873
9999261
9999519
9999409
9949301
9449074
9999921
9999951
9994970
99999S2
4999989
999U993
9999998
9999998
9999999
9999909
'1 0000000









0 6
0082940
0327123
0719102
1237982
1857448
2548177
3230302
4025G28
4759375
5461319
6116314
4714269
7249673
7720881
SI29287
8478393
8773116
9019110
9222177
9388418
9322991
9631017
9710934
9794661
9837569
987S327
9909944
9933821
99M798
3963205
9471109
9982356
4947407
9991376
9994053
9995418
9997251
999S157
994X774
9999195
9999475
9909661
4999733
•>
-------
                                                      Station Positioning
                                               Appendix B - Error Analysis
                                                              August  1986
Therefore:

                   c = 13.1/29.9 = 0.44

and                K = 30/29.9  = 1.003

Entering  Table B-2 for values K  =  1.0 and c = 0.4 gives  a probability of
62 percent.

     Equipment manufacturers quote accuracy  probabilities using various
terminology.  Most  commonly,  CEP and d    values are provided.  For equal
standard deviations (ox/ay  =  D,  the d *  is 1.2 times the CEP.  Thus,
quoting the  CEP would be more advantageous to a manufacturer than  quoting
a d    value.   However, when  system  accuracy is given  as 100 m (328 ft)
at fife5 2o level,  two clarifications are needed:  1)  whether absolute or
repeatable  accuracy is being  quoted, and 2)  whether   measures linear or
circular error.   If the error is  circular, the number describes an 86 percent
probability level,  whereas if  a 2d    is referred to, then the probability
of the positioning being within  the crrcle  is between  95 and 98 percent,
depending on  error ellipse  axis  lengths.  Therefore, unless the basis of
an accuracy  figure is very explicitly stated, it may be necessary to clarify
its meaning.  Only then can the accuracy performance of competitive systems
be compared.

     The above discussion is concerned  only with the accuracy of the positioning
method.  However, the drift of sampling equipment  away from the  vessel
(which  is usually manifested  as  a  wire angle) will  affect the size of the
area from which the sample  may have been collected.  The  distance the equipment
may drift  is,  in effect,  added  to the radial error of  the position fix.
The increase in  the probable area above that inherent in the positioning
method  is demonstrated in Figure  B-4.  This larger area  of probable sample
location is  the  total error associated with  the  location of a  sampling
station, and should be considered by the user in determining if the achievable
accuracy is  adequate for the purpose of the  study.

      The increase  in the probable sampling  location will  be a  function
of both the  wire angle and the depth of the  station.   The increase of the
sampling area with depth at various wire angles is exhibited  in Figure B-5.
The wire angle or equipment drift may not always  be  noticed or avoidable.
A  5  degree  wire angle  is  not  uncommon in shipboard sampling  and  anything
less than 2 to  3 degrees  may not be  noticeable.   Wire  angle may not be
avoidable and the worst-case  that may have to be accepted  during  a  study
should be used  in determining the probable sampling area.   The  probable
sampling areas  in Figure  B-5  have been calculated  by  adding wire  angle
errors to the radial error  (calculated from  Equation 2) for +2-m  (+6.6-ft)
accuracy positioning methods.

     The accuracy of the  positioning system chosen also  will affect the
probable area of  sample collection.  The less accurate the navigation method,

                                    B-9

-------
        VESSEL
        OFFSET
   WIRE
  ANGLE
                                                        RADIAL ERROR (RADIUS
                                                        OF PROBABLE LOCATION)
RADIUS OF PROBABLE SAMPLE COLLECTION AREA
    Figure  B-4.   Components of error  in  probable area  of  sample
                  collection.
                                   B-10

-------
    2500-1
    2000-
    1500-
r»
oc
    1000-
    500-


    300-

    200-

    100-

      0
             NOTE. BASED ON ± 2m ACCURACY AT A
                   90 DEGREE FIX CALCULATING THE
                   1 dm, AT 68% PROBABILITY
                         I
                        100
 I
200
 \
300
 I
360
                                  DEPTH (m)
  Figure B-5.   Effects  of wire angle on the  probable  (P = 0.68)
                area of  sample collection.
                                 B-ll

-------
                                                       Station Positioning
                                               Appendix B - Error  Analysis
                                                               August  1986


the greater  the resultant  area potentially sampled.  The effect  of depth
on the probable sampling area as a function  of the accuracy of the positioning
method  is  presented in  Figure B-6.   The 5 degree wire angle increases  the
probable sampling area significantly at depths below 100 m (328 ft).

     To ensure  that the preferred  positioning method will meet all  study
goals,  the  error associated  with sample  collection at  specific stations
should  be estimated. The  achievable accuracy can be determined by the radial
error utilizing the crossing angle and  standard deviations of the LOPs
expected over the entire study area.   Either Equation 2 or 3 can be  used
to calculate  radial  errors at specific probabilities.   Tables B-l or  B-2
can be  used  to determine probabilities  associated with a specific radial
error.   The wire angle error component  is  then  added to  the radial error.
The distance  added  to the radial error  (see Figure B-4) can be calculated by:

                             d = z(tane)                              (6)

where:   z = the station depth
        9 = the wire angle.

A  5 degree angle may be unavoidable and  should always be used to calculate
the wire angle error component.  Substituting into Equation 6 gives:

                             d = 0.09(z)                                (7)

This distance, added to  the  radial error,  will determine  the probable  area
of sample collection for  any station.
                                    B-12

-------
                   30.000 -,
                                                                           90.000 -i
                    20.000 -
                                                                           80.000 —
                 111
                 cc
                                                           ±20m
03
i
                    10,000-
DC
                                                                           70.000 —
                                                                           60.000
                                                                                                          1100m
                                                                                       NOTE BASED ON A 5 DEGREE WIRE ANGLE
                                                                                            AND 1 d,m> CALCULATED FROM A
                                                                                            90 DEGREE FIX (68% PROBABILITY)
                                                                400
                  100
                             I
                            200
                                                                                                               I
                                                                                                              300
 I
400
                                                                                                DEPTH (m)
                    Figure B-6.   Effects of navigational  accuracy on the probable  (P  = 0.68)  area of sample
                                  collection.

-------
                                                        Station Positioning
                                                Appendix B - Error Analysis
                                                                August 1986
                                APPENDIX B

                                REFERENCES
Bowditch, N.  1984.   American  practical navigator.  An epitome of navigation.
Defense Mapping Agency Hydrographic/Topographic Center, Washington, DC.

Mikhail, E.M., and  G. Gracie.   1981.  Analysis  and adjustment of  survey
measurements.  Van  Nostrand Reinhold  Co., New York, NY.  340 pp.
                                    B-14

-------
              APPENDIX C

POSITIONING METHODS AND CONSIDERATIONS
      FOR  SAMPLING  IN  PUGET  SOUND

-------
                                                       Station  Positioning
                                          Appendix C - Positioning  Methods
                                                               August  1986
                               APPENDIX C

                POSITIONING METHODS AND CONSIDERATIONS  FOR
                         SAMPLING IN PUGET  SOUND


     The various  navigational  positioning  techniques are listed in Table C-l.
This  information provides quick-reference review of the performance character-
istics  and  costs of methods representative of each class.  These methods
are grouped by maximum range to facilitate user comparison for the appropriate
sampling  sites within Puget Sound.  Optical methods are presented first,
followed by  electronic systems.

OPTICAL POSITIONING TECHNIQUES

Methods Available

     The traditional  optical  positioning method involves observations of
two horizontal  sextant angles between  three fixed shore  targets,  plotted
as a graphical  resection  using a three-arm protractor or station pointer.
Other optical  positioning methods are available  (Table  C-l), but only  this
and  the theodolite intersection method  are practical in more open waters
[150 m (492  ft) to 5 km (3.1 mi) from shore].  Because  the remaining optical
methods cited apply only  to river or harbor  surveys, at extremely short
ranges, or in very calm water, they are not useful over  the range of expected
conditions of sampling in Puget Sound and  are only briefly discussed below.

     When sampling stations are located relatively close to shore in industrial
areas, they can be identified by  relatively inexpensive  methods.   Use of
a  graduated tape or wire  is impractical  except in the case  of very short
distances offshore.  Intersecting ranges are used  when a  number of established
landmarks permit easy  selection of  multiple  ranges  that  intersect at the
desired  sampling point, and accuracy is not critical.  Range  line and uniform
speed  or angle  between  the line and an onshore target are more applicable
to hydrographic surveys than  discrete sampling  station  locations.   Vertical
angle  ranging requires  a  graduated  shore line target known  as a  subtense
bar.  A sextant is set so the individual marks on the  bar subtend  a given
angle  at desired distances from  shore.  This method  is limited to extremely
confined areas, relatively short distances, and calm waters.   The angle
and  stadia method was  used historically in  very calm waters, but can be
used at longer ranges.  This  method  requires vertical positioning of the
stadia (or rod)  on the  vessel  and  careful  measurement  of  angles to the
stadia hairs, enabling a  calculation of distance  from  the observer to the
rod.   A position  fix  is  made by measuring  the azimuth  from  a baseline of
known orientation.
                                    C-l

-------
    TABLE C-l. SUMMARY OF VESSEL POSITIONING METHODS

Close Range-Direct iup to 5 km (3.1 rrnj]
     Horizontal Sextant Angle Resection
     Theodolite Intersection from Shore
     Subtense Ranging by Vertical Angle
     Intersecting Ranges
     Range Line and Angle from Shore or Vessel
     Angles from Shore and Vessel
     Angles and Stadia or Distance from Shore
     Range Line and Uniform Speed
     Distance Line Ranging

Close Range-Indirect [up to 5 km (3.1 mi)j
     Laser
     Infrared Electromagnetic Distance

Short Range [up to 40 km (25 mi)J
     Variable Range Radar
     Microwave Electronic Positioning
     (300 MHz-300 GHz)

Medium Range [100-300 km (62-186 mijj
     Medium and High Frequency Electronic Positioning
     (300 KHz-300 MHz)

Long Range [to 2000 km (1243 mi)j
     Low Frequency Electronic Positioning
     (30-300 KHz)

Global Positioning
     Very Low Frequency, Satellite, Astronomical Observations
     (3-30 KHz)
                           C-2

-------
                                                       Station Positioning
                                          Appendix C - Positioning Methods
                                                              August 1986


     An optical  range  finder may be  used  to establish  the  predetermined
distances  to  permanent targets.   An optical range finder is used  by  simply
focusing  a  split-image on  the  target, enabling the slant distances to the
target to  be  read off the instrument  scale.  When  combined with  a careful
compass reading, this distance reading allows positioning of the sampling
vessel.  A  survey of accuracies claimed for commercially available instruments
suggests  that +3 m (9.8  ft) accuracy can be achieved within  approximately
100 m (328 ft)  of the sighted targets.  The Lietz Model 1200,  for example,
provides an  accuracy of +1 m (3.3 ft) at 100 m (328 ft).  Beyond this distance,
instrumental  errors increase rapidly.  For the instrument cited,  a  +9 m
(29.5  ft)  accuracy is quoted  at 300 m (984 ft).  The suggested U.S. list
prices of  optical range finders  vary  from $35  to $120 (Folk,  L., personal
communication).  Accuracies are affected by  the availability  of targets
of known position and  compass  readings.  Both of these  errors are  hard
to quantify,  making this method  unpredictable and of limited usefulness.

     Visual  lines have sometimes been used to establish a station position.
This method requires that a minimum of two objects are in alignment, enabling
the vessel to be placed on a common axis extending  to the vessel's position.
Simultaneous sighting on a second set  of at  least two objects places the
vessel  at the intersection of the  two common axes.  The accuracy of each
visual line is highly dependent  on the quality of the visual  range  (e.g.,
alignment),  the distance from the  alignment  objects to the vessel, and
the angle  between each  range.  Also,  the number of  visual fixes used affects
the magnitude of the positional  error. Although this technique is frequently
used for positioning single  sampling stations in bays and harbors or otherwise
where a number of convenient alignable targets can be selected,  the method
has  limitations for large  scale monitoring  programs.  Even when  a  desired
sampling  station has  been  accurately located using one of  the  previously
discussed  methods, and  at  that location at  least three aligned target-pairs
are  visible, it not likely that sufficient  alignment  target-pairs will
be present for all other desired locations.   Also, the  unpredictability
of  positional error each time a station is  reoccupied detracts  from the
value of this method.

     Because the techniques described  here are  inexpensive to implement
(as  are use of the  sextant resection  of theodolite intersection  methods),
they are  attractive to  small  studies.   However,  use of more sophisticated
and  less labor-dependent techniques may  be  achievable  at moderate costs
by  renting or leasing,  rather than  buying  such  equipment.  All such  options
should be considered, provided the recommended  absolute positional  accuracies
can  be achieved by  the  technique selected.

Theodolite Intersection

     Position of the sampling  vessel can be  established by two  onshore
observers who simultaneously measure the  angle between  a reference object
or  shore  traverse  and  the vessel (Figure  C-l).   A rod or  other aiming  point
is  normally erected  on  the  vessel.   Radios,  flags,  or lights   signal  the

                                    C-3

-------
Figure C-l.   Station positioning by theodolite intersection,
                            C-4

-------
                                                       Station Positioning
                                          Appendix  C  -  Positioning Methods
                                                              August 1986


moment at  which angle measurements  should be made.  A theodolite with an
accuracy  of +15  sec for single  angle  measurement,  intercept angles  near
45°, and a range of 5 km (3.1  mi), should yield a position error less than
+1 m (3.3 ft)  (Ingham 1975).   Characteristics of theodolites used for  such
measurements are summarized in Tables C-2 and C-3.

     Although accuracy of this method appears high, its use  in open waters
has several distinct disadvantages.  Complex arrangements usually are needed
to ensure  that  angles are measured  by two onshore  observers at the same
instant of  the desired fix.  Though  not a problem when the vessel  remains
on station for  a long period of time, it becomes a  problem when the vessel
is trawling.  Lines from  the  two theodolites  should  intersect at  nearly
right  angles.   As indicated in the  error analysis discussion, weak position
fixes or  corresponding large positional areas  of uncertainty result  when
the angles measured are  less  than 30° or more than 150°.  To avoid such
extreme angles,  onshore observers may have  to  move  frequently, which can
delay  a  predetermined sampling schedule.  Each of  the  onshore observer
stations must be  surveyed to maintain accuracy.  Finally,  as with all optical
methods,  target movement and  path interferences  (e.g.,  fog, heavy rain,
or heat waves) can confound the measurements.   In spite of these disadvan-
tages, the procedure offers relatively high accuracies  at  low capital cost
(although labor  cost can rapidly add up) and has been successfully  applied
in small-scale  coastal surveys during  favorable weather.  It  also is advan-
tageous in very  restrictive areas or in harbors.  It is therefore considered
a candidate method for very limited  monitoring  programs or  surveys.

Sextant Angle  Resection

     An offshore position can  be  fixed  by measuring  the  two  horizontal
angles between  1ines-of-sight to three identifiable  targets with known
positions. When a vessel is underway, the sextant angles  must be measured
simultaneously by two observers.  The measurement of the first angle  between
the center and one outside target allows determination of a circle of position
(COP) on  which the sampling vessel must lie  (Figure C-2).  For  example,
when  the measured angle  is  52°, the first circle of  position is  plotted
by subtracting this angle  from 90° and  drawing  lines seaward from the siting
targets  at the  resultant angle of 38°  from  the baseline.  These lines cross
at the center  of the  COP, which can then  be  drawn with a  compass.  This
procedure is  repeated using the center and  remaining targets,  resulting
in  the plot of  a second  COP.   In the example, the  second angle  is  67°,
requiring a plot of  radius lines at 23° from the  baseline.   The intersection
of the two position circles marks the  vessel's  location.

     Position fixes  are  normally plotted with  a  station pointer or  a three-
arm protractor (Figure  C-3).  The  two measured angles are  set  by moving
and  locking the two outer arms of the protractor,  which then is moved  over
a nautical chart until  the three arms  align  with the  preplotted  locations
of  the shore targets.   The  vessel's position  is  recorded at the center
of the protractor.   Because this  procedure  can be  implemented  in 10-15


                                   C-5

-------
TABLE C-2. SUMMARY OF VERNIER TRANSIT AND SCALE-READING THEODOLITE CHARACTERISTICS
COMPANY
Benchmark
Berger
Kern
leltz
Nikon
Pentax
Teledyne
Topcon
Warren -Knight
White
Mild
Zelss
MODEL
JENA 020
Bronze 65/45
Aston 67
Project 100
ST-1/6
ST-8/9
Kl-S/ST
BT20/10C
115
TSZOA/S6
NT-2S MK III
Schneider 700/400
GT-4B/6B
TH-60S/60E
OP 107/100-A.ZO
6-15
400U
AG-30B
TL-605E
10-2220/3200
TR-300
TR-303/303PM
T-307AT/309T
T-16
T-05
Th-42/43
VERNIER MICROMETER
OR SCALE DIVISION
20"
20V11
20"
r
1V20"
20"
30"
20"
r
r
r
20V11
20"
r
20"
20"
r
30"
r
2ovr
r
20"
r/20-
20"
20"
POSSIBLE
ESTIMATION
10"
20"/30"
20"
30"
30-/20"
10"
6"
10"
30"
20"/6"
0.2'
6-/20"
6"
20"
20"
15"
20"
10" /30"
6"/10"
12"
10"
10"
U.S. SUGGESTED
LIST PRICE*
S2495
S2125/3850
$1835
$1100
$699/1499
$1699/1899
$3895/3995
$1695/1995
$795
$2495/3695
$3195
$1235/650
$1695/1895
$2500/1895
$1295/1185
$1350
$650
$1595
$2100
$2695/1295
S749
$1879/1995
$2695/2350
$1895
$3950
$3950
•January 1985
                                      C-6

-------
                          TABLE C-3. SUMMARY OF MICROMETER AND DIGITIZED  THEODOLITE CHARACTERISTICS
o
I
COMPANY
Benchmark
Kern
Lletz
Nikon
Pentax
David White
Tope on
Wild
ZeUs
MODEL
JENA 010A
JENA 015B
DKM2-A/T
IKH3/DKM3-A
DT-20E
TM6/10E/20H
TM-1A
HT-1/5A
NT-4D/30/2D
TH-20D/10D/06D
TH-01N
TH 10-20/10
T.308AT/208AT
TH 10-1
OT-20
TL-20DE/10QE/6DE
T-O/T-1
T-2/T-3
T-2000
Th-2
ITh-2
VERNIER MICROMETER
OR SCALE DIVISION
1"
6"
1"
0.5"
20" Accuracy
6V10-/20"
1"
zovi"
6"/10"/20"
20V10V6"
r
20V10"
20-/W
r
LCD Readout 20*
20V10V6-
20"/6"
1V0.2"
0.5'
1"
0.6"
POSSIBLE
ESTIMATION
o.i-
1"
0.1"
0.1"
10" Display
2-/5VIO"
0.1"
6V0.25"
l"/2"/«"
5"/2"/r
0.5"
2"/l"
10"/5"
0.1"
10"/5"/3'
20"/3"
o.r/o.1-
0.1"
-
U.S. SUGGESTED
LIST PRICE*
14295
S3495
15890
SI 2295
S2995
13995/3695/3295
t5995
SI 895/5950
$3695/3595/3095
12850/3500/3690
14900
J2850/3500
12995/3495
(5500
S269S
t 2859/3500/3690
$3995
$5995/1199$
$11995
$5950
116600
                 January 1985

-------
o
I
00
                   Figure C-2.  Station fix using position circles.

-------
           VERNIER
                                                   MOVABLE
                                                    SIDE
                                                    ARMS
    INDEX ERROR
    ADJUSTMENT
        SCREWS
   SETTING
   SCREW
Figure C-3.   Three-arm  protraccor for  sextant resections,
                               C-9

-------
                                                      Station Positioning
                                          Appendix C - Positioning  Methods
                                                              August  1986
sec with  some experience,  it  commonly is used  in  hydrographic surveys where
moving vessel positions are needed.   To  minimize the parallax  error, the
two sextant  operators should stand  as close as possible when making the
measurements.  Sextant angles  can  routinely be measured  to approximately
1 min of  arc or better,  depending upon the instrument quality and  operator
ability.   Within 5 km (3.1 mi) of  shore stations and at  acceptable COP
crossing  angles, the resulting  accuracy in  position is 1 part in  2,500,
or about  +2 m (6.6 ft) (Ingham 1975).

     Sextant angle  resection has been the most widely used  positioning
technique in coastal surveying. This is due to  its relatively high accuracy,
ease  of  implementation, and  nominal  cost of the sextants and the  three-
arm protractor.  Also, no  shore  party is required.  The  procedure does
have some limiting factors, however.  Range is  ultimately limited  by visibility
and by the sizes, elevation,  and  placement of the shore targets.  Also,
it is imperative to follow procedures in locating targets to avoid  indeterminate
or weak fixes.   For example, a fix cannot be obtained when  the vessel and
all three shore targets lie on a common circle (called the danger circle).
This can be avoided by aligning the  shore targets along  a  straight line,
which causes the radius of  the  danger circle to become  infinite.   Another
recommended practice  to assure strong fixes  is  to place targets so that
intercepted  angles fall between 30° and 140° (ideally between 45° and 60°),
thereby maintaining large position circle cut  angles.  Shore targets also
may be  placed  on a  curve  convex to  the observer, with the middle target
nearest the sampling  vessel (Figure C-4).  Alternately, targets may lie
on a  curve concave  to the vessel, provided that the anticipated positions
are within the  triangle  formed by them, they  are virtually equidistance
from  the vessel, the observed angles  are not less than  60°, or the vessel
is well outside  the circumscribing circle (Clark 1951; Davis et  al. 1966).

     At the  limits  of visibility, sextant angle fixes are  likely to  be
weak  because  the angles are small.  In such cases, large  positional  errors
can  be caused  by small  errors  in  the angles themselves (Umbach  1976).
This  problem  can be  partially offset by using  a  telescope mounted  on the
sextant.   Sextants  must  be  in perfect adjustment and angles  must be read
with  extreme  care.  Also, the sextant  must  be  of  superior quality so  that
the angles may  be read to the necessary  precision. When working at locations
near  shore,  the sum of  the  two  angles can  approach 180°, with one angle
often very large and  the other very small.   Under  such conditions,  care
must  be taken to mark the two angles  simultaneously if  the vessel is moving,
or to make several measurements   if the vessel is on  station  because  the
angles rapidly  changed with slight vessel movements.

      Split-fixes (no common center object) may  be taken  when  a three-point
 fix is not possible.   The vessel  position is at the intersection  of  two
angle loci.   A  fix  is  considered  strong when  the  intersection angle is
greater than 45°.   While shoreline tangents can be  taken when no other
objects are  available these  fixes are inaccurate  in most cases.   Split
 fixes and shoreline tangents  should be  used only when absolutely necessary.


                                   C-10

-------
DANGER
 CIRCLE
                                                                            Adapted  from Ingham 1975
        Figure C-4.  Shore target  locations  to  avoid  the  danger  circle.

-------
                                                       Station Positioning
                                          Appendix  C  -  Positioning Methods
                                                              August 1986
They are  inefficient due to the  required recording procedures and plotting
time,  and  they cannot generally be entered into  automatic data processing
and plotting  systems.

     Sextants are classified as vernier or micrometer drum types, the latter
preferred  by  most users.  A well-constructed metal  marine  sextant is capable
of measuring angles with  an  instrument error not exceeding 10 sec or 0.1
min of arc (Bowditch 1984).   However, as indicated earlier, positional
error will be  highly dependent  on operator ibility.   For accurate work,
a sextant having an arc radius of 162 mm (6.4 in) or more  should be selected.
Characteristics  of representative sextants are  presented in Table C-4.
Sextants must be adjusted  prior to the start of  the survey and verified
at the  conclusion or at  least  once a week, whichever  is more frequent.
Any index  correction should be recorded in  the  survey log.  Procedures
for sextant adjustment are provided by Umbach (1976) and Bowditch (1984).

     Considering the achievable accuracy when the  double  horizontal  sextant
angle method  is properly implemented, this procedure offers an inexpensive
candidate positioning method for limited surveys.  However, the method
may be ineffective during poor visibility or periods of heavy weather  when
monitoring or seasonal survey work may nonetheless  be required. Also important
is the need to construct, survey, and maintain a sufficient  number of properly
located shore  targets to  provide unambiguous positioning.  Overall, the
method has merit when the cost of an electronic  positioning system  cannot
be justified.

Variable Range Radar

     While technically not an optical  positioning technique,  obtaining
ranges from a variable range radar (VRR) consists of two optical estimations
by the  user  (target location and bearing of  radar reflection) and will
be addressed  herein.  A position can be fixed  by  measuring the distances
to three  targets on the  radar screen that can  be accurately identified
on a map.  A third  fix will reduce the chance of error and  increase accuracy.
A variable range marker measures the distance to the object (as identified
by its radar  reflection).  This distance then  is  drawn  with a compass as
a line  of position (LOP)  on  the nautical chart.  The  intersection of the
three LOPs marks the vessel's position.

     Most  commercial vessels are equipped with radar for safety and navigation.
Coastal  vessel radar usually have ranges  from 26 to 116 km   (16 to  72  mi).
A variable range marker (VRM), whether  built-in or added onto the existing
on-board radar, removes  a  large portion of operator error in estimating
distances.   For  positioning  at  accuracies less than 300 m (984 ft),  a VRM
almost always is needed.  Range accuracies with the VRM are usually  1 to
2 percent of the range  scale,  or +25 m (82  ft)  at 0.5 km (0.3 mi), the
smallest scale.  Accuracy decreases as the range scale is  increased.   Bearing
                                   C-12

-------
                      TABLE C-4. MARINE  SEXTANT CHARACTERISTICS
SPECIFICATIONS
Arc Range
Instrument
Accuracy
Vernier
Scale
Arc Radius (on)
Telescope
Brightness
Index Mirror (on)
Horizon Mirror Ota. (on)
Illumination
Frew Mtl.
Arc Mtl.
Case
U.S. Suggested
C. PLATH
NAVSTAR
CLASSIC
-5 to «I2S°
< ilO"
0.2'
162
4X40
6X30

57X41
57
Arc A Drum
Brass/Al loy
Brass/Alloy
Mahogany
t!270
NAVSTAR
PROFESSIONAL
-5 to .125°
< 410"
0.2'
162
4X40

57X30
57X30
Yes
Makrolon
Al. Alloy
Plastic
J965
TAMAVA
SPICA
-S to «I25°
< MO"
0.2'
162
««0(7°)0
7X35(6.5°)
100/25
57X42
57
Arc 1 Drum
Aluminum
Bronze
Plastic
S919
TAMAVA
JUPITER
(UNIVISION)
-5 to M25°
<*«•
0.2'
162
4X40(7°»
7X35(6.5°)
100/25
57X42
57
Arc ft Drum
Aluminum
Bronze
Plastic
S71S
TAMAYA
VENUS
-5 to *125°
< 18"
0.2'
138
3X26
75
45X30
45
No
Aluminum
Bronze
Hood
SSBS
UEEMSft
PLATH
-5 to <
< ±10"
0.2'
162
4X40
6X30

57X41
57

.125"







Arc A Drum
Brass
Brass


Mahogany
11674

List Price (1/B5)

-------
                                                       Station Positioning
                                          Appendix C - Positioning  Methods
                                                               August  1986


accuracy  is  usually less  than 1°.   Characteristics of representative  VRRs
are presented  in  Table C-5.  Generally, analog systems have better resolution
than digital  systems, but are not as  versatile.

     Each position  fix will rely on the resolution and identification of
the radar reflection.   Resolution of  the  target position will  change with
the range scale.  Reflections will depend upon  target position and alignment.
The location  of  the  reflection is not always  easy to identify.  Placement
of  the cursor away from the actual  reflection surface will  introduce error
into the  fix.  Estimation of the radar target  location in relation to the
chart's  mapped  structures is important. Misidentification  of a reflection
source will  result  in  plotting a position  at the wrong coordinate, but
will  not  affect repeatable  accuracy if the  same target is used.  Certain
regions, such  as  sloped  headlands and  tidal flats, give inaccurate reflection
because  it  is impossible  to relate the reflection to a map location.  The
most accurate radar  range fixes are obtained from solid reflections between
0.16  and  6.4 km (0.1  and 4 mi).  This keeps the range scale low, resulting
in accuracies greater than  +40 m (131 ft) and avoids erroneous readings
caused from very close  reflections.

     Positioning vessels with  VRR  should  provide sufficient increases in
accuracy over  standard radar for most sampling purposes.  Repeatable accuracies
can  be within +20 m (66  ft).   Position fixes from VRR are  still available
in almost any type  of  weather.  Most vessels already  are equipped with
VRR  and  other radar can  add variable  range  markers for $1,000 to  $2,000.
The newer digital  systems, priced in  the $5,000 to $10,000 range,  offers
multilevel  processing for better target pickup and provides map plotting
ability directly on  the screen.

ELECTRONIC POSITIONING  TECHNIQUES

     Electronic positioning methods  use  the transmission of electromagnetic
(EM) waves from  two or more shore  stations  and a vessel  transmitter to
define a vessel's location.  The systems  are  based on the ability to predict
variations in EM wave travel velocity as  a function of travel path. Position
is  determined by measuring differences in signal arrival times (range-range
mode) or by comparing  the phases of received signals to that of the transmitted
signal (hyperbolic ranging).

     At their respective maximum ranges,  electronic positioning methods
have higher accuracies  than visual methods.   Measurements can be obtained
regardless  of weather and visibility conditions, and operating ranges are
typically much greater  than for  optical methods.  Range can be extended
to  50-100 km (31-62  mi) simply  by elevating  antennae until signal attenuation
becomes a limiting factor.  Shore stations need not be attended, minimizing
personnel requirements.   Positional readouts  are  available as distances  or
coordinates,  rather  than wavelengths or time delays,  and  deck  units are
                                   C-14

-------
                               TABLE C-5. VARIABLE RANGE RADAR  (VRR)  SYSTEM CHARACTERISTICS
r>
i
Systems
DECCA
Racal DECCA Marine
Redmond. UA
(206) 885-4713
Furuno U.S.A.
San Francisco, CA
(41S) 873-9393

KODEN/SI-TEX
Norwell, NA
(617) 871-6223
Raytheon Marine
Seattle, UA
(206) 285-6843
Model
R0170 VRM/VP3
R0170 BT


FR-360 MKII

FR-810
FR-1011
T-100


3604
3610

Type
Digital
Digital


Analog

Digital
Analog
Digital


Analog
Analog

Variable
Range Markers
1 (add-on)
1


add-on

2
1
1


1
1

Range
(k«)
77
77


58

116
77
26


58
116

Nominal Accuracy
Range (m) Angle
M
+30 at 0.4 km


+25 at 0.5 km

+36 at 0.5 km
725 at 0.5 km
+22 at 0.5 km


.
-

+0.50
+0.50


<+lo

<+lo
<+lo
+0.50


+10
+10

U.S.
Suggested
List Pricea
$3.990
16.000


$4.400
without VRM
$7.400
$8.300
$2.200


$4.900
$5.400

              December, 1985.

-------
                                                       Station Positioning
                                          Appendix C - Positioning Methods
                                                              August 1986
usually compatible with data  processing and automatic plotting equipment.
Position information is continuous,  enabling maintenance of a  desired location
by dynamic  positioning or by traversing along a predetermined  path.

     The short-range systems of primary importance in survey  and monitoring
programs are compact, lightweight,  durable, easily calibrated,  and relatively
stable.  Disadvantages of electronic  systems can include cost,  particularly
for smaller program requirements, the inconvenience of orienting  shore
and on-board units, vandalism of shore stations, and unknown  signal propagation
effects (although this should not be a problem over  the relatively  short
distances  to survey points  in Puget Sound).  As discussed  later, costs
can be minimized by sharing the expense among  multiple users, by  leasing
equipment  during  survey periods, or by contracting for  survey personnel
and equipment.

System Classifications

     Electronic positioning systems often are classified by  range capability,
which depends largely upon  propagation characteristics of  the operating
signal.   Band width and  signal  power also  influence range  capability.
Electronic  positioning methods are herein categorized as short-range, medium-
range, and  long-range systems.  Although short-range systems are emphasized,
other categories also are  examined  because  systems such as  Loran-C are
frequently used in parts  of Puget Sound.  Satellite navigation systems
also are presented because their prices are  declining and capabilities
(i.e., coverage and repetitive access) are expected to increase. Categories
of electronic positioning systems,  including operating frequencies,  wavelengths,
ranges, and representative  commercially available equipment, are  listed
in Table C-6.

     Short-range  [0-40 km (0-25  mi)]  microwave systems  are  portable and
best suited  for use  in  the range-range mode.  Medium-range  systems [to
150  km (93 mi)] also are  transportable, although  components usually are
bigger  and  heavier.  They are effective in  either  the range-range or hyperbolic
mode.   Long-range [to 2,000 km (1,243 mi)]  and  global systems  transmit
from permanently installed, widely dispersed  shore stations  or satellites
for multiuser operation.

Comparative Absolute Accuracies

     Although it is difficult to specify the positional accuracies achievable
by instruments in each category, some generalizations can be made.   Whereas
the  optical methods discussed can provide accuracies of +2 m  (6.6 ft)  for
ranges up  to 5 km  (3.1 mi)  offshore, short-range electronic  positioning
systems may provide accuracies of  +1-3 m  (3.3-9.8 ft)  for  ranges up  to
40 km (25 mi) from  shore stations.  Comparable medium-range system  accuracies
are  +5.0  m (16 ft) up to  150 km (93 mi).  Long-range systems typically
have accuracies  of  +50-100 m  (164-328 ft) within 350  km  (217 mi)  of shore
stations,  and more  than 100 m (328 ft) at greater ranges.


                                   C-16

-------
   TABLE C-6.  ELECTRONIC  POSITIONING  SYSTEM  CATEGORIES
Category
           Representative
Range          Systems
Very long range           >2,000 km

  Very low frequency
  Satellite
Long range                0-2,000 km

  Low frequency

Medium range              0-150 km

  Medium-high frequency



Short range

  Radar                   0-100 km
  Microwave               0-40 km
           OMEGA
           TRANSIT (NAVSAT)
           GEOSTAR
           NAVSTAR GPS
           SERIES
           AERO SERVICE GPS

           LORAN-C
           VIEUNAV
           LAMBDA

           SYLEDIS
           RAYDIST TRAK IV
           HYPER-FIX
           ARGO OM-54
           HYOROTRAC
           DECCA
           FURNO U.S.A.
           KODEN/SI-TEX
           RAYTHEON MARINE

           TRISPONGER
           MINIRANGER
           MICRO-FIX
           HYDROFLEX
           AUTOTAPE
           AZTRAC
           POLARFIX
           ARTEMIS
                           C-17

-------
                                                       Station Positioning
                                          Appendix C - Positioning Methods
                                                              August 1986
Operating Modes

     There  are two principal  radio navigation  system operating  modes:
the two-range  (or range-range) mode and  the hyperbolic mode.   Some  systems
incorporate the advantages  of each in a combination mode.  These three
modes are presented in Figure C-5.   In. the range-range mode,  position fixing
is accomplished by measuring  the extremely small time intervals required
for EM signals to travel  from an on-board  master transmitter  to  one or
more onshore slave stations, and back (Figure C-5a).  For a known  propagation
velocity, the  time interval  is converted to a  distance  (range)  from the
slave, defining a single  circle of position on which the  vessel may  lie.
The intersection of two or more such circles  (based on signal  returns  from
two or more slave stations) results  in a  position fix.  This operating
mode usually is restricted to a single user,  although  single  side-band
techniques  have been employed to allow multiuser operation (Ingham 1975).
Lane width (distance between points of zero signal phase)  remains  constant
regardless  of distance from the system  baseline.  The lane width  resolution
does not  decrease at increasing ranges  from  the baseline,  as is  the  case
in the hyperbolic positioning mode.

     In the  hyperbolic mode, the on-board receiver detects the phase difference
of signals arriving from multiple shore-based transmitters.  Lines of constant
phase  between master and  slave transmitters  form a hyperbolic  pattern of
position  lines (Figure C-5b).  By measuring  the phase difference between
arriving  signals, the vessel can be located along one of the position hyper-
bolas.  Adding a second master-slave transmitter pair superimposes another
hyperbolic  pattern, resulting in a grid network with  pattern  crossings
(Figure C-5c).  Measurement of the signal phase  difference  from the second
transmitter  pair allows vessel positioning on the second pattern, and therefore
"unambiguous" location at the applicable grid crossing point.  Phase differences
usually are  resolved to 1/100 of the lane width.  Resolution  of the hyperbolic
system matches that of  a  range-range system only along  the  master-slave
transmitter baseline.   Because of lane widening for increasing  range  from
the baseline,  the system  resolution  decreases  with vessel  distance  from
the master-slave transmitter baseline.  As indicated earlier,  the angle-
of-cut of arriving  signals also affects the  magnitude of position error.
In hyperbolic systems,  extending the baseline length improves  cut-angles
of the arriving signals and decreases lane  spreading  (Ingham 1975).

Short-Range  Systems

Electronic Distance-Measuring  Instruments—

     A position  fix  is obtainable with two electronic distance-measuring
instruments  (EDMI), or  one EDMI with an  angle measurement by theodolite
or  sextant.  Distance-measuring instruments  that are either electro-optical
(e.g.,  laser) or electronic  (e.g., microwave)  are discussed herein.  Electro-
optical  and microwave  distance-measuring  devices  are extensively used  in


                                   C-18

-------
n
i
                  RANGE-RANGE
               GRID CONFIGURATION
B.
     HYPERBOLIC PATTERN
                                                                                     LANE
                                                                                    WIDTH
C.       HYPERBOLIC
    GRID CONFIGURATION
                Figure  C-5.  Operating modes for radio  navigation systems

-------
                                                       Station Positioning
                                          Appendix C - Positioning Methods
                                                              August 1986


land-based  surveying.  The EDMI  master generates  a  carrier signal which
is directed  toward a reflector  (in  the case of  light  beams) or a  repeater
(for microwaves).  The light or microwave beam is modulated  at two or three
different  frequencies, usually under the control  of a precision  quartz
crystal oscillator.  A phase  comparison of the incoming and outgoing beams
enables accurate distance determinations.

     The EDMI is  relatively new  in the  surveying  field.  The geodimeter
of the early 1950s, which  used a modulated  light beam, was replaced in
the late 1950s by  the Tellurometer, which used a modulated microwave  signal.
This improvement  increased the range and allowed operation in  moderate
rain,  fog,  and  darkness.   Newer  EDMIs have shorter ranges, but due to the
use of solid state electronics, are much more compact, less power-intensive,
and easier  to  read.  The  newest  EDMIs use highly  coherent laser light,
have longer  ranges,  require even less power,  are portable, and  are  easy
to operate. The so-called "total station" consists of a theodolite for
measuring  angles and an EDMI for measuring distances, with outputs  recorded
on magnetic  or paper tape for subsequent analyses.  Under favorable conditions,
EDMI range capabilities are 1.6 km  (1 mi) for  light-based  systems, 80 km
(50 mi) for  laser  systems, and  150 km (93 mi) for microwave systems.

     Properly adjusted and calibrated,  an  EDMI has few sources  of error.
Ground wave  reflection  can cause  error when  measurements are made  over
water because reflected signals result  in faulty distances due  to the longer
path lengths. The  swing, or cyclic manner, in which reflections are recorded
must be correctly  interpreted.   At very close range, EDMI accuracy  is limited
by a constant of  uncertainty.  Beyond 0.5-1.0  km (0.3-0.6 mi), accuracies
of  1  part in 25,000 are easily achieved.   If  meteorological conditions
over the signal  path  are  sufficiently well-known, accuracies of  1  part
in  100,000  can  be achieved (Moffitt and Bouchard  1982).  Characteristics
of representative short- and  long-range electro-optical and microwave distance
measuring devices  are  presented  in Table C-7.  As indicated, instruments
with a range of  25 km  (16 mi) can  be used to  measure distances  to  within
5  cm  (2 in), while  shorter-distance devices [to  5 km (3.1 mi)]  are accurate
to 1.5-3.0 cm (0.6-1.2  In)].

      It is  apparent,  therefore, that accuracies  achievable with electronic
distance measurement devices are more than adequate to meet the positioning
requirements for surveys and monitoring programs.   In fact, it  is  the angle-
measuring devices used with  EDMIs that limit accuracy, not the EDMI itself.
Probably the major disadvantage of EDMIs is the  continual motion and resultant
misalignment of  the reflector (in electro-optical  systems).  Use of  microwave
patterns eases directivity  requirements.

Total  Stations—

     An electronic tachymeter, commonly referred  to as a total station,
is an  instrument for determining the distance,  bearing, and  elevation of
a  distant object.  In  coastal surveys, it  is a shore station  instrument

                                   C-20

-------
TABLE C-7. ELECTRONIC DISTANCE MEASURING INSTRUMENTS
COMPANY
Benchmark
Orlando. FL
(305)281-5000
Ceodineter. Inc.
Nova to. CA
(415)677.1256


Kern Instruments
Brewtler. NY
(914)279-5095
Keuffel 1 Esser. Co.
Morrlstown. NJ
(201)285-5000
The Heti Company
Overland Park. KS
(913)492*4900
« Electronics. Inc.
Littleton. CO
(303)795-20<0
Nikon. Inc.
Garden City. NY
(516)222-0200
Pentai. Corp.
Englewood, Co
(303)733-1101
Teludlst. Inc.
Mastic Beach. NY
(516)399-5843
Tope on Instrument Corp.
Paramus, NJ
(201)261-9450
U114 Heerbrugg
Farmingdale. NY
(516)293-7400


MODEL
Surveyor II 1-1
Ceodlmeter 14-A
Geodineter 112/122
Geodlmeter 220
OH 503
Ranger V-A
(HeNe Later)
PulteRanger
RED 2A/2L
MK-I1I, MX- Ml VS
NO 31
PK-81
Tellenat CMW20
(microwave)
DM- S3
Cltatlon-450
01*41 Dlstonat
01-20 Omomat
RANGE (m)
SINGLE PRISM
TRIPLE PRISM
MAX|MUM( prisms)
1600
3000
3500(6)
6000
8000
15000
2500
3600
6000(8/16)
1600
2400
3200(8)
2000
3500
4500(7)
8000
16000
25000
1000
3000
2000/3800
2800/5000
	 /7000(9)
1600
3000
4000
1900
3200
1400
2000
25000
2000
2500
2900(9)
1600
2300
4000(11)
2500
3500
7000(11)
6000
7000
14000(11)
ACCURACY
(MSE)
i(5 nm • 5 ppm)
1(5 mm • 3 ppm)
*(5 m * 3 ppm)
«(5 nn * 3 ppm)
2(3 m « 2 ppm)
2(5 mm « 2 Ppm)
2(30 W » 150 ppm)
2(5 nm * 5 ppm)
2(5 mm « 2 ppm)
2(5 mm • 5 ppm)
*(5 mm • 5 ppm)
2(5 mm •» 3 ppm)
1(5 mm * 5 ppm)
2(5 mm * 5 ppm)
1(5 nm * 5 ppm)
1(3 mm « 1 ppm)
U.S. SUGGESTED
LIST PRICE
S3.49S
SI 1.300
S6. 250(112}
110.950(122)
SB. 850
S8.995
S20.561
17.500
S4.69S(?A)
$5,950(111)
S7.9SO(II1-VS)
S5.885
S4.790
$16,500
$5.390
$3,995
18,995
$14.995
                      C-21

-------
                                                       Station Positioning
                                          Appendix C - Positioning Methods
                                                              August 1986


used to  sight the survey vessel  reflectors, enabling positional information
to be recorded onshore for subsequent communication to the  vessel  operator.
In a manual  station, the same telescope optics (co-axial) are  used to measure
both distance and angles.   They  basically are  theodolites with  built-in
EDMI units.  With such manually operated units, slope reduction of distances
is done by optically reading the  vertical angle and keying  it  into  a  built-
in or  hand-held calculator  (McDonnell, Jr. 1983).  A semiautomatic total
station contains a vertical  angle sensor for  automatic slope  reduction
of distances, while horizontal angles are optically read.  With an automatic
station,  both horizontal and vertical  angles  are electronically  read  for
use with slope distances in a data collector or internal computer.  A theodolite
with a mount-on EDMI  usually is not classified as a  total station.   An
exception is the modular total station, where the design objective is flexi-
bility of future additional  equipment.  Such units usually are  designed
around an electronic (digitized) theodolite  such as the Kern El.  Many
total stations are designed to make full use of hand-held calculators (e.g.,
the  HP-41CV)  for data storage, computations, access to control  registers,
testing, calibration, and orthogonal offset determinations. Most manufacturers
offer  optional  data  collectors that serve as  electronic  supplements to
field books. This permits a convenient interface with a computer  and remote
transmission of  data using an acoustic modem.

     Characteristics  of representative total   stations  are presented in
Table C-8.  As  indicated,  range is dependent  upon the  number  of  prisms
available for  signal reflection.  Such prisms  are directional  (i.e., must
be pointed towards the shore station), as opposed to omnidirectional  prism
arrangements used for range-azimuth navigation systems.  Directional prisms
cost approximately $300 for a pair and $500 for  a set of  three.   Although
systems  requiring 9  to  11  prisms are expensive, they allow measurements
regardless of vessel orientation.  Some manufacturers report  displaying
angles or accuracies to  1  sec  or less.  However, experienced  surveyors
know that this  level of  accuracy requires careful or  repeated  pointings
at good targets.

     For  surveys  and monitoring programs,  single station capability is
attractive.   Setup and calibration efforts  are minimized, and the logistics
of  station  movement  are  much simpler than  with a multi-station net.    A
total  station  can  be used on other  projects  when not  in  use for periodic
monitoring.  The $8,000  to $30,000 price  range is competitive with the
microwave positioning  systems  ($40,000 to  $100,000),  and achievable accuracies
in  both  range  and angle  are more than adequate.   Instrument  capabilities
and  costs are  reported in the free, bimonthly journal  Point of Beginning
[P.O.B. Publishing Company, Wayne, MI (313-729-8400)].

Microwave Navigation  Systems-

     Short-range electronic  positioning  systems  generally operate at microwave
frequencies that  limit  the  system  ranges to "radio  1ine-of-sight."  Typically,
such systems are effective between 25 and  100  km (16  and 62 mi) offshore,

                                   C-22

-------
                             TABLE C-8. TOTAL STATION CHARACTERISTICS

COMPANY


Carl Zeiss, Inc.
Thornwood, NY
(914)848-1600


MODEL TYPE RANGE*
(km)
L M H
RMS3 Send -Auto - 1.5?
- 2.0b
- 3.0e -
Elta 3 Automatic 1.6

PRISMS


1
3
9
3
ACCURACY
RANGE ANGLE


+5-10 mm + Zppm +2"


±10 mn + Zppm +2"

u.s. S'jciiri:
LIST PRICE

58,260


S18.7Z5




Geodlmeter, Inc.
Nova to. CA
(415)883-2367

Kern Instruments
Brewster, NY
(914)279-5095
'et:
•Hand Park. KS
,13)492-4900
MK electronics
Littleton. CO
(303)795-2060

Nikon Instruments
Garden City, NJ
(516)222-0200

Pentax
Englwood, CO
(303)773-1101


Topcon
Paramus, NJ
(201)261-9450



Wild Heerbrug
Farmlngdale, NY
(516(293-7400





Elta 3


Elta 46R
140



E1/DM503
E2/DM503

SMO-3
SOM-3ER
SET-10
MK-III
MK-IV
MK-HYDRO

NTD-4

NTD-1

PX-10D


PX-060

GTS-2B


ET-1


T2000+
DI4L or
DI5

T2000+
DI20


Automatic


Automatic
Automatic



Automatic


Manual
Semi -Auto
Automatic
Automatic



Manual

Automatic

Manual


Manual

Manual


Automatic


Automatic



Automatic



-
.
-
-
1.2
1.8
2.4
3.6
1.5
2.0
2.4
.
.
-
_
.
-

m
.
-
-
m
.

.
-
.
•
-
.
.
-
1.2
1.5
1.7
1.8
2.0
2.3
2.6
2.7
1.6
2.5
3.0
2.0
2.2
3.0
3.8
4.8
2.5
3.5
4.5
0.8
1.4
-
3.0
4.0
3 or 4

1.2
1.8
1.2
1.8
1.4
1.7

1.4
1.7
1.4
2.0
2.6
1.4
2.0
2.6
2.5
3.5
4.5
5.5
6.0
7.0
8.0
9.0
.
.
-
-
3.0
4.0
5.5
6.0
3.0
4.5
5.5
1.2
1.8
2.5+
_
.
-

1.6
2.3
1.6
2.3
m
.

.
-
1.7
2.4
3.0
1.7
2.4
3.0
3.5
5.0
6.0
7.0
9.0
11.0
13.0
14.0
3
6
18
3
1
3
6
8
1
3
7
1
3
9
3
3
3

1
3
1
3
1
3

1
3
1
3
9
1
3
9
1
3
7
11
1
3
7
11
±10 mm


±10 mm
l(5mni «



1(3 mm
1(3 mm

1(5 mm


1(5 mm
+ 2ppm


+ 2ppm
• 5 ppm)



* 2 ppm)
* 2 ppffl)

* 5 ppm)


» 2 ppm)
±2"


±3"
12"



12"









10.5"

10"
5"

6"

digital
direct


Stationary
1(20 mm + 5 ppm)
Moving
1(5 mm

1(5 mm


* 5 ppm)

+ 5 ppm)

!(5ran + 5 ppm)


±(5 mm

1(5 mm


1(5 mm


1(5 mm
1(3 m


l(3mn <





+ 5 ppm)

+ 5 ppm)


+ 5 ppm)


* 5 ppm)
+ 2 ppm)


" 1 ppm)




6"
3"
I"

ID-
S'1
2"
6"
1"
6"


2"


Q.51




digital
S18.7Z5


$12,320
$19.950



$19,175
$22.375

$8,800
$10,800
$11.000
$12,950
$14,950
$16.950

$8,595
subdivision


digital
direct
estimation
digital
$15.985

$8,450


$8,765
estimation






i



0.5"






$7,990


$14,250


521,000
$25,000


$30,000



* Atmospheric visibility:  a. low - hazy, 5km
b.  medium • clear, 15km




      C-23
c.  high • very clear, 30km

-------
                                                       Station  Positioning
                                          Appendix C - Positioning Methods
                                                               August  1986


depending  on  antenna heights  and power outputs.   Position measurements
are indirect (i.e., by timing the travel of  multiple pulsed  signals  from
a master to two or more remote stations  and back; alternately,  phase differences
between arriving  signals can be measured).   Available systems  operate in
the range-range mode, the  hyperbolic mode, or both.  The position  fix is
defined by  the intersection  point of two  position circles or  hyperbolic
constant-phase  lines.  Because microwave  systems have nominal  accuracies
of +1-3 m (+3.3-9.8  ft)  from very short  ranges to  25-40  km (16-25  mi),
they provide  adequate positioning capability for survey work and monitoring
programs.  Potential  limiting factors  include problems with shore station
security,  and signal interference in industrial areas or in the vicinity
of radar-intensive military installations.   Characteristics of representative
microwave navigation  systems are summarized  in Table C-9.

     Trisponder—The Del  Norte Trisponder  is an X-band (8,800-9,500  MHz)
positioning system composed of a digital distance-measuring  unit (DDMU),
a master station (usually  on the vessel),  and two remote  stations located
at known geographic  positions.  Each  station  is a  combined transmitter
and receiver.  The master station antenna  is  omnidirectional and each  remote
station has  a directional antenna.  Distances  to remote stations are observed
on the  DDMU  using the  range-range mode.   A time-sharing feature  allows
up to eight users.

     The manufacturer quotes  a typical  range accuracy of +1 m (+3.3  ft),
with an instrument resolution of  0.1 m (0.3  ft).   In The Hydrographic Manual.
Umbach  (1976) cites  a  range error for this system of +3 m (+10 ft)  with
good field conditions  based on the Trisponder Basic Operation  Manual published
in  1974.  Also cited were  tests conducted by  the National  Ocean Survey
(date unknown),  which indicated  that  temporal electronic  drift may cause
measurement variations.   Recal ibration  is suggested over a measured  base
line after every 200  h of operation.   However,  the  tests were conducted
on a Model  202 Trisponder Surveyor System for which the manufacturer claimed
a resolution of  3 m  (10  ft) and  a  positional accuracy  of +3 m (+10  ft),
after field calibration.

     The currently available microwave system operates with Model 217/218E
transponders and Model 520 or  542 DDMUs.  The Model 261 transponder,  usable
as  a master or  remote, has a  5-km (3.1-mi) line-of-sight  range.  The Model
217/218E transponders operate  up to 80 km (50 mi)  from shore due to  higher
output power.   The transponders  are designed for  use  with either of the DDMU
models.  The Model  520 will  collect four ranges and display  two.  The Model
542  interrogates four remotes,  outputs four sets  of  data,  and also provides
a positioning guidance capability.  The cost of a complete  system, including
a Model  520 DDMU, a  master  and  two  remote 217E/218E transponders,  and antenna,
is $40,000 (Buchanan,  C., personal  communication).   For  the  additional guidance
capability of the Model  542  DDMU,  the system cost  is $44,500.  A less expensive
"black box" version of the  DDMU  (Model  562)  is available  for use with an
existing shipboard  computer  system, allowing  more  than  eight users.  Cost
of  the Model 562 and other  required components is $39,500.

                                   C-24

-------
                               TABLE C-9. SHORT-RANGE POSITIONING SYSTEM CHARACTERISTICS
r>
i
SYSTEM
TRIPSONDER (520/540 ODMU)
DEL NORTE Technology. Inc
Euie*>. TX
(817)267-3541
FALCON 484 Mini-Ranger
Motorola. Inc.
Tempe. AZ
(602)897-4376
MICRO-FIX
Racal OCCCA Survey. Inc.
Canarlllo. CA
(805)987.8080
HVDROFLEX
Teludist. Inc.
Mastic Beach. NY
(516)399-5843
AUTOTAPE ON-40A/OH-43
Cubic Western Data
San Oiego. CA
(619)268-3100
RANGE FREQUENCY
(kai) ("HO
5 9320-9500(261)
80 9329-9500(21 7E)
80 B800-9000(2I8E)
37 5410-5600



80 54JO


100 3000

ISO 2900-3100


NOMINAL USER
ABSOLUTE CAPABILITY
ACCURACY
1> up to 8
(•ore
optional)

i? up to 20



±1 up to 16


*l .5 single
pTos 3x10 cull i user
di stance (•) option

^0.5 single
~1 : 100 ,000 range


ANIENNA
Vessel Shore
Station
320°?

360°H
2S°¥



360°H „
20°,30°V


360°H
18HV

3CO°H
IOBV


87°/18ql?M
66/50°»

7fljM
15 V


900H



I8°¥.60BH
30°-I20°H
IO°V


COST
140.00C

139,300



$43.000


S63.500

190.000


             •January 1985

-------
                                                      Station Positioning
                                          Appendix  C - Positioning Methods
                                                              August 1986
     Falcon 484 Mini-Ranger—The Motorola Falcon 484  is a C-band (5,410-5,600
MHz) microwave ranging and  positioning system  that  operates in the  two-
range mode from  100 m (328 ft)  to  40 km (25 mi).   The system consists of
a vessel  receiver-transmitter  assembly with an ommidirectional antenna,
a range  console, and shore-based radar  transponders with directional antennae.
Pulsed radar from the survey vessel  transmitter interrogates radar transponder
reference stations located at geographically known points.  Elapsed time
between  transmitted interrogations  and the reply  from  each transponder
is used  as the  basis for  range  determinations.  Two ranges are used for
trilateral positioning. If three or four ranges are available, range residuals,
sum of  squared  residuals,  and error  circle radius  data are output for the
least squares position solution.   The manufacturer claims  a  range accuracy
of +2 m  (+6.6 ft).  Up to 20 users can  operate in the same area.

     The current cost of a basic Falcon 484 system, including a range processor
control  display, receiver/transmitter with ommidirectional antenna,  two
reference stations with directional  antenna, miscellaneous cables, and
manuals  is $39,300  (Jolly, J.,  personal communication).   In an effort to
eliminate one of  the two  major  problems with most microwave positioning
systems  (i.e., magneton failures  and  need to  service  beacons), Motorola
currently is developing a solid-state beacon using a gun diode.  The associated
increase in  long-term reliability also will  result in  decreased  range.
However, range  capability should remain adequate  for  positioning needs
in Puget Sound.   Costs of  the modified transmitters  were not available
at time  of publication.

     Micro-Fix—The Racal Survey  Micro-Fix is a range-range  microwave posi-
tioning  system.   With 1ine-of-sight  to shore-based transmitters,  it is
capable  of operating up to 80 km (50 mi) offshore.  The  system normally
operates at 5,480 MHz, with options at 5,520  and  5,560 MHz.  The  master
station  can  interrogate up to eight  remote stations (from  a  possible 32),
with each remote transmitter/receiver  unit  (T/R)  preset  to recognize its
own distinctive station code.   Up  to four separate station  groups can be
deployed in the same  area without interstation interference.   Multiuser
capability  allows a maximum  of  16  users for each deployed chain.  Master
and remote units are  interchangeable.

     The basic system consists  of  a master  station with  a  Control Measurement
Unit (CMU),  a T/R unit, and two tripod-mounted remote T/R  stations.  The
vessel's master station  interrogates the remote  stations  sequentially,
triggering reply pulses  received  by the master T/R and processed  by the
CMU to display the  corrected ranges.   The CMU capabilities include automatic
and continuous self-calibration,  track  guidance,  plotter drive, x-y  conversion
(full  spheroid and  multi-range  solution),  and slant range correction.
Nominal  accuracy is stated  by the  manufacturer to be +1 m (+3.3  ft).  The
cost of a basic Micro-Fix system is $43,000 (Harris, E.,  personal  communica-
tion), including training.
                                   C-26

-------
                                                        Station Positioning
                                           Appendix  C  -  Positioning Methods
                                                               August 1986


     The  manufacturer  is  developing the  capability to use the system in
 a hyperbolic mode, a combination range-range/hyperbolic mode, or a range/azimuth
 mode, but this capability is not yet available.  The  system also uses circular
 polarization techniques  to avoid reflective (water surface)  signal cancellation
 nulls, thereby eliminating  the  need for a second antenna  on the vessel. This
 has been accomplished  by an antenna  design that  prevents  signal entrance
 from reflective angles.

     Hydrof 1 ex—Hydrof 1 ex is a  short-range to medium-range microwave navigation
'system designed for survey applications where a  high degree of accuracy
 is  required  in  fixing or tracking  a moving vessel.   The system operates
 at  frequencies of 2,920 to 3,300  MHz and has a  range of  100 m to  100 km
 (328  ft to 62 mi).   It consists of  a  master unit controlled by an HP85
 computer with a customized  software  package,  an  omnidirectional antenna,
 and connecting  cables.  Each remote unit consists  of a  transceiver and
 either a large range or fan-beam antenna for mounting on  a customer-provided
 tripod.  Accuracy is claimed to be  1  m (3.3 ft) +3xlO~6D, where D  is the
 distance in meters.  The system can be operated in either a two- or  three-
 range  mode.  Although single-user is  the normal operating mode, a multiuser
 option  is available.  The cost of a master and two remote  stations is $63,500
 (Baker, W., personal communication).

     Autotape DM-40A/DM-43—The  Cubic Western Autotape  is an S-band  range-
 range microwave positioning system that operates at ranges  of up to  150 km
 (93 mi).   System  components  include a shipboard  interrogator and range
 responders at each fixed onshore station.  The OM-43  is capable of working
 with three geographic sites.  Range information is  computed by comparing
 the phase shift of the modulated signal  transmitted between the interrogator
 and responder phase unit to an interrogator reference  signal.  An Automatic
 Position Computing System  (APCS)  is available  for  steering information,
 real  time analog  plot of the vessel's track, and magnetic data recording.
 The system does not accommodate multiple users.   The  manufacturer  claims
 a  range accuracy  of  0.5 m plus  1:100,000 times the  range distance.  This
 error  is due  to  internal  random  errors,  systematic  errors, temperature
 variation bias errors, signal  strength, component aging, and initial calibration
 errors.  External errors are  said  to  far exceed internal  noise, systematic,
 and bias errors.   Index of refraction  error, which can  approach 5 m (16.4 ft)
 at  100  km (62 mi), is usually  small  enough  at  short  ranges to be  ignored.
 Multipath rms errors (dependent on  orientation of reflective objects  near
 and behind the interrogator omnidirectional  antenna) have been  observed
 from 0  to 3 m (0 to 9.8 ft).   The  manufacturer states that  internal averaging
 plus external data smoothing will  reduce the effect on  Autotape  to  a small
 fraction of  a meter, provided the antenna  is  moving.  For this  reason,
 the system  is best suited to  applications where the interrogator is on
 a moving vessel.

     The cost of a basic Autotape  system including an interrogator, two range
 responders, and associated  antennae  is  $90,000.   The same system  with  the DM-43
 and three shore  stations is $124,000 (Hempel,  C., personal  communication).

                                    C-27

-------
                                                       Station Positioning
                                          Appendix C - Positioning  Methods
                                                               August  1986
Medium-Range  Systems

     Systems  in  this category  typically operate  in  the medium- to  high-
frequency bands  (i.e.,  1.5-400 MHz), achieving greater  ranges using EM
waves  that  propagate around  the earth's  surface.  Positional  accuracies
of medium-range systems vary from a few meters near the base  line to tens
of meters  at the system's range limits (Ingham  1975).  Medium-range systems
must be used with caution in inland waters  due  to the severe landmass attenu-
ation and water-land  interface effects.  Such  effects usually are manifested
as large calibration variations within a limited survey area.  Characteristics
of representative medium-range electronic  position fixing systems are summarized
in Table C-10.  Because of the limited availability, high capital costs,
and logistical problems posed by the large-sized land stations, these systems
are not discussed further herein.  Other  systems offer similar accuracies
at reduced  costs  and are  less time  intensive for station setup.  Various
semi-permanent systems in Table C-10 are  discussed in detail  in Tetra Tech
(1986).

Long-Range Systems

     Long-range and global navigation  systems generally operate  at  low
(30-300 kHz)  or very low (less than  30 kHz)  frequencies.   As in  the case
of medium-range  systems, EM waves at such frequencies travel  for very long
distances, typically limited by transmitter power.   Onshore station chains
usually are  permanent,  for use with an appropriate vessel  receiver  and
published hyperbolic lattice charts (Ingham 1975).  Achievable accuracies
typically are much  lower than those of shorter-range systems because long-
range systems  are designed for  general  navigation rather than  accurate
positioning.  Satellite  navigation  systems,  which operate at much higher
frequencies, afford global  coverage at much higher  accuracies.  Characteristics
of selected long-range navigation systems are  presented in  Table C-ll.

Loran-C—

     Loran,  an acronym  for  long-range navigation,  is a pulsed low-frequency
electronic navigation system that operates at  90-110 kHz  in the hyperbolic
mode.   Loran-C  receivers match cycles  to measure  time differences between
arriving master (M) and  secondary  (W, X, Y,  Z)  signals,  which are  pulse-
and  phase-coded  to enable source identification  (Panshin 1979).  The usec
arrival time  differences are displayed and can be plotted on  a special
Loran-C latticed chart as  lines-of-position.  Fully  automatic Loran-C receivers
simultaneously process  signals  from two master-secondary station  pairs,
displaying LOP information for  tracking.

     Range  capability varies  because Loran-C stations  radiate peak powers
of 250 kw-2 I*W.  Due to the use of  low  frequencies and  large  baseline distances
[i.e.,  1,850  km (1,150 mi)  or more],  Loran-C can provide positional  information
of reasonable  accuracy out to 2,225  km (1,380 mi) with sky waves (Maloney

                                   C-28

-------
                              TABLE C-10. MEDIUM-RANGE POSITIONING SYSTEM CHARACTERISTICS
SYSTEMS
SYLEDIS
Sercel , Inc.
Houston, TX
(713)492-6688
RAYDIST
Hastings-Raydist
Hampton, VA
<-» (804)723-6531
IN)
*° HYPERFIX
Racal DECCA survey
Camarillo, CA
(805)987-8080
ARGO DM-54
Cubic Western Data
San Diego, CA
(619)268-3100
HYDROTRAC
Odom Offshore Surveys
Baton Rouge, LA
(504)769-3051
RANGE
(km)
300



276 night
740 day



700 day
250 night


740 day
370 night


460 day
230 night


FREQUENCY
(MHz)
420-450
406-434


1.5-2.5




1.6-3.4



1.6-2.0



1.6-4.0



NOMINAL
ABSOLUTE
ACCURACY
±1 LOS
±3 2X LOS


±3




5-10
0.01 lane
display
resolution
4-5
0.05 lane
display
resolution
2-40
0.01 lane
display
resolution
USER ANTENNA
CAPABILITY VESSEL SHORE
Up to 4 Not Provided
(Range-range)
Unlimited
(Hyperbolic)
4/net Omnidirectional
(Range-range)
Unlimited
(Hyperbolic)

(Range-range) Omnidirectional
Unlimited
(Hyperbolic)

Up to 12 Omnidirectional
(Range-range)
Unlimited
(Hyperbolic)
Single Omnidirectional
(Range-range)
Unlimited
(Hyperbolic)
COST*
$ 51,350



$102,200




$113,000



$188,000



$100,000



*January 1985

-------
TABLE C-ll. LONG-RANGE  POSITIONING SYSTEM CHARACTERISTICS
SYSTEM
LORAN-C
Multiple Receiver
Manufacturers
LAMBDA
DECCA Survey, Ltd.
England
ONEGA
Multiple Receiver
Manufacturers
VIEUNAV
Navigation Sciences
Bethesda, MO
(301)952-5225
TRANSIT
Multiple Receiver
Manufacturers
GEOSTAR
Geostar Corporation
Princeton. ID
(609)452-1130
NAVSTAR/GPS
Multtple Receiver
Manufacturers
SCRIES
IS! AC. Inc.
Pasadena. CA
(818)793-6130
ACROSERVICE
Houston. TX
(713)784-5800
RANGE
(k»)
2500 day
I8SO Night
400-800
Global
Vre
Coastal
Areas
Global
U.S. Land I
Coastal Areas
Global
Global
Global
FREQUENCY
(KH()
90-110
100-200
10-14
90-110
150 ft 400 MHi
5117-7075
1618-2492
1575 MHi
1228 MHi
1575 MHz
1228 MHi
1575 HHJ
1228 MHi
NOMINAL
ABSOLU1E
ACCURACY
(•»)
185-460
15-90 repeat able
137 day
730 night
1800 day
3700 night
110
90. 1 pass/1 freq.
37-46.1 pass/2 freq.
3-S.Multiple pass
and 2 freq.
1-7
40.C/A.CEP
8-9. P. CEP
< 1 B dlff.
Mode
< 1 n
(1/2 hour)
USER
CAPABILITY
Unlimited
Single 2- range
Unlimited -hyperbolic
Unlimited
Multiple
Unllnlted
Unlimited
P-Ntlitary
C/A Conwrctal
Foreign
Unlimited
Multiple
COST
$1.000-2,000
Call Agent
$4.000-10.000
$40.000 and
$2,000 annual fee
$2.500-10,000 basic
$30.000-45.000 elaborate
Transceiver $450 and
link Rental $10-30/m>.
(Estimated)
$10.000-140.000 Initial
$1.000 eventually
$287.000
$235.000

-------
                                                       Station Positioning
                                          Appendix C - Positioning Methods
                                                               August  1986


1978).  Range achievable at a particular station  is dependent upon transmitter
power,  receiver  sensitivity, noise or interference levels,  and signal  path
losses  (Canadian Coast Guard 1981).

     At best, the  absolute accuracy of Loran-C in  normal operating  mode
over short distances using the ground wave  varies from 185 to 460 m (0.11  to
0.29 mi),  whereas repeatable accuracy varies from 15 to 90 m (49 to 295
ft) depending on the vessel's location within  a  given coverage area  (Dungan
1979;  U.S.  Coast Guard  1974).  Achieving the  short-range accuracies cited
above requires proper  installation,  maintenance, and operation of high-
quality  equipment (Canadian  Coast  Guard 1981).  Higher accuracies can  be
obtained  by operating  in  a differential mode (i.e., with a supplemental
receiver at a surveyed  site onshore which transmits corrections or offsets
to the  survey vessel).  Available  equipment  varies from  simple receivers
and  indicators  to  fully  automated receivers with self-tracking capabilities
that can interface with a  vessel's computer. The cost of a Loran-C receiver,
excluding the antenna, varies from less than $1,000 to $2,500-plus.

     Although Loran-C frequently is used in  Puget Sound for sampling and
monitoring, this application has potential  problems.  Of particular  concern
when operating  in  areas  such as Puget Sound are time and spatial variations
in the  Loran-C signals,  and signal  interferences that  prevent operating
in desired survey areas.   Inland location [up  to 160 km (100 mi)] of Loran-C
chains  requires  overland  signal transmission.  This results in phase  shifts
that are difficult to predict.  Such shifts can cause an erroneous position
location fix.  There are also anomalies associated with land-water interfaces
and  large structures such as bridges and tall buildings.  Crossing-a   ^s
also can  widely  vary  from one geographic area  to another.   In some  ca^s,
lines of  position are almost parallel, making  an accurate fix very difficult.
Noise and interference (e.g.,  from engines  and other electronic equipment)
also can  be disruptive, but most Loran-C receivers are equipped with factory-set
or tunable notch filters  to minimize  such problems.

     In  an  effort  to improve navigational  capability using  the Loran-C
system, the U.S. Coast Guard completed a one-time survey of the  east and
west U.S. coasts  in which Loran-C  positions were compared with those from
a calibrated microwave system.  Corrections were obtained for the  Defense
Mapping  Agency  nautical  charts, whose LOPs were based on theoretical trans-
mission over water paths (Ryan, R., personal communication).   These corrections
do  not include  seasonal  or diurnal  signal effects.   The  land transmission-
path effect  (known as  the additional  secondary phase  factor)  currently
is  under evaluation.   The results  of a recently completed multi-year West
Coast Stability  Study, which extended from San  Diego  to  Vancouver  Island,
indicate that  the repeatability  of  Loran-C  in  Puget  Sound  is significantly
better than other parts of the  country.  The annual  variation  in  signals
were on  the  order of 0.2 usec (Figure C-6).  The corresponding 95  percent
confidence ellipses  and 2 drms  positional  repeatabilities  at Neah  Bay are
+40 m  (131 ft).   As the  System Area Monitor at Whidbey Island  is  approached,
seasonal  stabilities  should  improve,  resulting in  improved  positional repeat-

                                   C-31

-------
r-i
i
(A>
                  .3

                  .2

                   1

                  00

                  -.1

                  -2

                  -3
                        31
Mean
Signs
                        '  • .  I
                                         Neah Bay
                                 Yankee
                                                                              Neah  Bay
                                                                                           Zulu
m^i****^**^''*^ 1^11**!^
                                                       .4

                                                       3

                                                       2

                                                       1

                                                      00

                                                      -1
                            61
                           28798 863
                           .041
                                    121
                                         151   181
                                                 211   241  271  301
                                                                 331
                                                                      361
                                         Julian Day 4
                                            1984
                                                                                 31
Mean
Signs
                                                                                     61
                                                   41893.826
                                                   .059
                                                            121   151   181   211
                                                                Julian Day 4
                                                                   1984
                                                                                                              241   271  301   331
                                                                                                                               361
                     Figure C-6.   Seasonal  variations  in  Loran-C signals  at Neah Bay,  Washington 5990.

-------
             Station Positioning
Appendix  C  -  Positioning Methods
                    August 1986
                                                              August 1986


ability.   Based on the study and the  propagation model  developed, the Coast
Guard can  estimate the error  ellipse and 2 drms  error circles for most
locations within Puget Sound (Slagle,  D., personal communication).

     Due to differences in the  path conditions, measured time differences
(i.e., difference in arrival times of two simultaneously transmitted signals)
often are  different than  theoretically predicted.   In fact,  the deviation
varies depending upon the  receiver  location within the reception area.
So-called  spatial  variation or grid warpage  can  be removed by applying
corrections based on measurements at a nearby site that has been accurately
surveyed.   Such a  procedure was followed by  the  U.S.  Coast Guard during
an extensive survey of Puget Sound's major  ship traffic lanes.  The  purpose
was to  establish accurately  located way points  to which the vessel can
be navigated.   Following  prescribed  turning  instructions at a given way
point,  the vessel proceeds to the next way point, and repeats  the procedure
until  the destination is approached.   Given an absolute accuracy requirement
of +40  m (131 ft), the Coast  Guard  conducted  simultaneous measurements
at thousands of locations  within the Sound  with a Motorola Mini  Ranger
and a Loran-C receiver.   As a result, time differences at geodetically
known way points have been published and data for many interim  track  points
have  been  archived.  Thus,  for these points, a spatial correction can be
made.   To maximize the  absolute and  repeatable accuracy at a given way
point  requires input of  an  additional correction factor to compensate for
the daily and seasonal signal variations earlier addressed.

     For purposes  of locating sampling stations, it is possible to  utilize
the way point data to develop a correction  for other nearby locations based
on interpolation of data  from the nearest four way point stations  (Gazely,
L., personal communication).   The accuracy to which  the desired time differences
can be  calculated and  therefore the site located is dependent on distance
from the way points or track points to the  desired location.

     Metro completed a similar  study with fixes  from  a microwave system
to plot  the  corresponding Loran-C coordinates within Elliott Bay and  surrounding
areas.  Absolute accuracies  within  this mapped  region are increased to
40-100 m (131-328 ft) with repeatable accuracies substantially better  (Sturgill,
D., personal communication).   Metro and other groups update this  Loran-C
chart periodically, with a  new version expected during  the  first half of
1986.   Because repeatable accuracies of Loran  are acceptable for the monitoring
objectives, calibration of  Loran-C positions  is a reasonable alternative
for  long-term monitoring  projects by reducing costs incurred with  renting
microwave   systems  (to a  single expense).   If  the study  site  is  within or
near  the calibrated Loran-C stations set up  by Metro or the Coast Guard,
then interpolation  from  the nearest stations  may be adequate  for  accuracy
requirements.

     Both  temporal  and  spatial variances  from a predicted navigation system
grid can be  substantially  reduced or eliminated  by operating in a differential
mode  (U.S. Department  of Defense and U.S. Department of Navigation 1984).

                                   C-33

-------
                                                       Station  Positioning
                                          Appendix C - Positioning Methods
                                                               August 1986


A facility  (in this case  a  calibrated Loran-C  receiver) may  be located
at a fixed  point within an area of interest.  Loran-C signals are observed
in real  time and compared with signals predicted for the known position.
The differences between the observed signal  and predicted signal  is  transmitted
to users as  a "differential correction"  to upgrade the precision and performance
of the user's receiver  processor.  For Loran-C, the serving  radius for
correction  transmission may be up to 320 km (200 mi).  The U.S.  Coast Guard
studied the  ability of the differential  Loran-C to meet the  8-20 m (26-66
ft) accuracy requirement  of U.S. harbors and harbor approaches.  For the
Seattle area, the Coast Guard Double Range Difference MOD 2 Model predicts
that  a  20  m (66 ft) 2 drms  absolute position  accuracy  is feasible when
operating in a differential mode (Doughty and May  1985).  Tests using  Least
Squared Error, Alpha-Beta  Filter, and Linear Regression Model  approaches
to predict  differential  Loran-C time difference offsets  indicate that a
consistent  accuracy within  10 m (33  ft) is  not an unreasonable goal for
the particular differential  Loran-C system examined on the Thames  River
(Bruckner 1985).

     Radio  frequency (RF) interference was evident during the West Coast
study, both nearby and within the Loran-C band  of 90-110 kHz (Blizard and
Slagle 1985).  RF interference was noticeably observed at the Tacoma Washington
Harbor Monitor Site, and directly affected the Coast Guard's data  acquisition
and collection.   During the 16 months of operation at this site, a significant
amount of the data was  considered  of  poor quality, attributable to the
intermittent U.S. Navy transmissions  at 76.3  kHz.  Apparently  the strength
of the signal is so strong that it caused the 100  kHz tuned Loran-C coupler
to oscillate.  Conversations with area users  indicated that  many other
types of receivers also experienced  similar  problems,  thereby limiting
the use of Loran-C  in  the Tacoma  area.  The extent to which  the problem
is experienced further away from  Tacoma will  be  receiver-dependent.  The
U.S. Coast Guard found that use of notch filters  in  the Hood Canal/Bremerton
area eliminated  the  problem.   Identifing  the limit of  the interference
area  would require  transits away  from the  source area  while attempting
to notch out the interference, a task not included in the Coast Guard  study.

     Other areas with signal  interference include  relatively small  intermittent
sources in Elliott Bay and  at Seola Beach near  Sea-Tac Airport (Figure  C-7).
A large region of northern Puget  Sound often  is  affected apparently because
of another U.S. Navy  transmitter.  This source  precludes reception from
most  sets north of Everett  (in an  area with  normally good Loran reception).

      Two additional  sources  of interference  found in the West Coast  study
were  transmitter switching  and chain control effects.   The AN/FPN-44A tube
transmitters in the U.S.  West  Coast Chain  (9940) located  at Fallen,  Nevada
(Master); Middeltown, California  (X-Ray); and  Searchlight,  Nevada  (Yankee)
are  switched every  14  days for  routine maintenance,  resulting  in a  signal
offset  for approximately 28 days.  T,he resultant  40 nanosecond error/offset
                                   C-34

-------
LARGE AREA REGULARLY AFFECTED
 BY POSSIBLY A SINGLE POWERFUL
                     SOURCE
  OCCASIONAL SHORT-TERM INTER-
        FERENCE IN ELLIOTT BAY
   SEOLA BEACH TO PT. WILLIAMS
      POSSIBLE AIRPORT SOURCE
                                                 NOTE: MOST OF THE SOUTH SOUND
                                                 ALSO EXHIBITS INTERFERENCE FROM
                                                 A SINGLE SOURCE NEAR TACOMA
  Figure  C-7.   Regions of LORAN-C signal  interference in  Puget Sound.
                                  C-35

-------
                                                       Station Positioning
                                          Appendix  C  -  Positioning Methods
                                                              August 1986


would most  likely be detected  and  applied in the time difference to the
latitude/longitude conversion  algorithm.  The Canadian West Coast  Chain
(5990)  would  most likely exhibit  similar offsets.

     Also apparent  in  the study was  the  effect  of moving an Alpha-1 site
receiver that maintains the Control  Standard Time Differences.  When the
9940 Whiskey A-l Monitor was moved in December 1984, a  positional grid
shift of 300 nanoseconds occurred.   For a typical  gradient  of 305 m (1,000 ft)
per usec,  this is equivalent  to a 91.4 m (300  ft)  change in position.
Thus,  applying Loran-C  in  a  repeatable mode, the  user would have  found
his "past"  position moved by more than 91.4 m (300 ft).   In fact, depending
on crossing  angle, the difference in  positions  could even be more.   Thus,
users  in  the repeatable mode must  monitor and  compensate any changes in
the U.S. West Coast or Canadian  Chain (whichever is used) when they occur.

     In summary,  the accuracies  of Loran-C vary depending upon the location.
Prior knowledge of the area reception is required for an adequate determination
of its  useability.   Achievable accuracies have to  be acceptable to the
sampling/monitoring objective for the  system to  be  considered.  However,
because of  its relatively  low  cost,  ease of installation/ operation, and
reliability in known areas over the range  of conditions, Loran-C is a candidate
system to be considered.

Viewnav—

     To improve upon the positioning  accuracy of standard Loran-C receivers,
Navigation Sciences has developed Viewnav, an interactive computer  system
that uses differential Loran-C to position a vessel with a claimed repeatable
accuracy of  4.6 m (15 ft).  Absolute  accuracy of the system  is on the  order
of +10  m (+32.8 ft) at  the  90  percent confidence level, and +5 m  (+16.4
ft) at the 20  percent confidence level  (Newcomber, K., personal communication).
Loran-C offsets are obtained by interrogating onshore monitors established
by the company.  In addition,  a land-based microwave system is  used to
calibrate a vessel's initial position  or track.  The system is particularly
effective in ports and  harbor areas  where large buildings or the land-water
interface may alter Loran-C readings.

     A  distinctive  feature of  the  system is an electronic display  of the
survey area based on digitized nautical charts.  As the vessel moves, its
display position moves relative to depth contours and land boundaries.
Other waterborne radar  images in the area also  are indicated.

     A  full  system costs approximately $40,000 with a  supplementary  annual
service fee  of  $2,000  for chart corrections  and equipment maintenance.
The  base price  includes  a mainframe Ai-M16  computer with 512KB of main
memory, a flexible  disk drive with 1MB capacity, and a Winchester hard
disk  of 10MB capacity.   The basic system provides 5MB of chart  storage,
                                   C-36

-------
                                                       Station Positioning
                                           Appendix C - Positioning Methods
                                                               August 1986


which  equates to approximately 650 charts  depending on  scale selected.
Additional  charts  can be stored on floppy disks.  The manufacturer expects
charts and  Loran monitors throughout U.S. coastal areas by 1986.

     The system could  provide improved accuracies over Loran-C, especially
in areas of Puget  Sound  that have  not  been mapped for  Loran.  The  shore
stations require setup  time,  as does  the initial  position  calibration,
making  this  system  more attractive to long-term monitoring programs.  Loran-C
reception limitations again cause limited coverage of the Sound.

Lambda and  OMEGA—

     These  two  long-range systems are offshore  positioning systems that
do not provide adequate accuracies and/or coverage  for  station positioning
within Puget Sound.   Interested parties are referred to Tetra Tech (1986)
for more information on these systems.

Transit (Navsat)—

     The U.S.  Navy Navigation Satellite System (originally Project Transit)
consists of  a group of satellites in 106-min circular  polar orbits at altitudes
of approximately 1,411  km (877 mi).  The system also includes ground tracking
stations, a computing center, an injection station, U.S.  Naval  Observatory
time  signals, and vessel  receivers and computers.  Positional measurements
are based on  the  Doppler frequency shift that occurs  when the  relative
distance between the  satellite transmitter and vessel  receiver changes
(i.e.,  frequency increase upon closure and frequency decrease upon separation).
Provided the satellite  orbits are accurately known, it  is possible to locate
the receiver.  The nature of  the Doppler shift depends upon the exact location
of  the receiver  relative  to the satellite path (Maloney 1978).  The system
operates at frequencies of  150 and 400 MHz so  that ionspheric corrections
can  be made through  signal  comparison techniques.   The vessel's position
is determined based on  "known" orbital positions during  satellite  passage
and measured frequency  shifts.

     As originally designed,  at least one satellite  would  be within line-
of-site every 35 to 100  min.  However, at U.S. East  and  West  Coast  latitudes,
the  acceptable fix window is approximately every 90  min (Driscoll, C.,
personal communication).  This is caused, in part,  by  the requirement that
a satellite's maximum  altitude be between 15°  and 75° before a fix is considered
valid.   Another problem  occurs when two satellites being  tracked  have approxi-
mately the same  closest approach,  whereupon it  becomes difficult to know
which  one  is  monitored.  Typically, each satellite  provides  four  fixes
a day  on two successive orbits spaced by 12 h.  Because one satellite currently
is  inoperative and another  has weak batteries,  it may take longer (e.g.,
several hours at the  equator)  to  gain a  valid  fix (Booda  1984).

     A static position  fix  with Transit  using single-channel equipment
can be made with an accuracy of  approximately  90 m (295 ft).   Dual-channel


                                   C-37

-------
                                                       Station Positioning
                                          Appendix C - Positioning Methods
                                                              August 1986
receivers  improve single-pass  accuracy to 37-46  m (121-151 ft) (Hoeber
1981;  Maloney  1978).  With multiple  passes, an rms accuracy of 3-5  m  (9.8-
16.4  ft)  is  claimed by some  equipment manufacturers.   Using  translocation
(i.e., simultaneous measurements  at  two fixed  locations, one of which is
accurately known), estimated positional measurement accuracies are in the
range  of 0.5-1.5 m (1.6-4.9  ft) (Oriscoll, C.,  personal communication;
Moyer,  C.,  personal  communication)%  Magnavox  has developed a satellite
surveying program (MAGNET) that is  reportedly capable of simultaneous  location
of up to  10  sites with a point-to-point accuracy of 30 cm (1  ft) or better.
However, this system is designed for coordinated geodetic  studies  over
wide  areas using 3 to 10 MX  1505  receivers and recording data from 10 or
more satellite passes (Anonymous  1981).

     Transit  receiver  costs range from $2,500 to $10,000 for basic single
frequency units (Murphy, W., personal communication).  More elaborate multiple-
channel  systems, sometimes  in  combination with OMEGA, range in cost  from
$30,000 to $52,000 (Jolly, J., personal communication; Driscoll, C.,  personal
communication).

     Use of the  Transit system generally is not appropriate for  sampling
and monitoring programs because of the inaccuracy of the  system  for  fixes
from  either a moving vessel  or one occupying  stations for a relatively
short time.  A fix must be based  on  a  single  pass.  With  satellite passes
at 1-h  or 2-h intervals, multiple-pass data acquisition  is impractical.
Therefore, only the best single-pass accuracy of  37-46 m  (121-151  ft) can
be achieved.   Since  translocation is  designed for two fixed  stations rather
than one on a moving sampling vessel, and  since MAGNET applies  to  land
surveying only,  neither method will  provide a significant increase in accuracy
during the limited  time the vessel will be  at  any one  station.   However,
the largest disadvantage of the system  still is the  time constraint between
fixes, which does not make it appropriate  for surveys using expensive  ship
time.

GEOSTAR—

     GEOSTAR  is  a pulse radio  transmission system.  Just approved by the
FCC,  it will provide satellite information  for  positions within the  continental
U.S.  and its coastal waters by 1987.  Three geosynchronous satellites  (and
a  fourth as backup) will orbit the earth  at 37,000 km (22,991  mi) at  70°,
100°, and 130° W longitude.  System components include transceivers, satellites,
and computers at a ground center.  The  links  between  the ground station
and the satellites  will  operate at 5,117-5,183 MHz and 6,533 MHz,  while
user-satellite links  will be at 1,618 and  2,492 MHz  (Whalen  1984).  Should
a  satellite fail, the backup would be moved  into  a proper orbit by telemetry
command from the ground computer facility.

     The  user will  send a command through the  transceiver, which  relays
the message through  the satellites  to  a central  computer  at  the ground
center, reportedly  in  less than 1 sec.   The signal-arrival times from each

                                   C-38

-------
                                                       Station Positioning
                                          Appendix C -  Positioning Methods
                                                              August 1986


satellite  are used by the  ground computer to calculate the position of
the specially  coded transceiver.   The  information is then transmitted  back
to the  satellites and relayed  back  to the transceiver  in  a similar amount
of time.

     GEOSTAR  will  enable a typical single-shot positioning error of +2-7 m
(6-23 ft),  according to the  developer.  When  needed, accuracies  down to
+ 1 m  (3.3  ft) reportedly can be achieved with two-way  interaction, signal
analysis,  and averaging.   Users at a known elevation  (e.g., sea level)
will  have  greater accuracy due to much smaller  geometrical dilution of
precision  where only  two  (rather than three)  coordinates are  required.
Continuous operations in a differential mode also should  enable correction
inputs for such  errors as  ionospheric delays,  satellite  position drift,
and drifts in  satellite electronic delays.

     System designers estimate that when operable the cost  of  a basic hand-
held transceiver with a typewriter keyboard and LCD display  will  be  less
than  $1,000.   A monthly  service charge in the  range  of $10 to  $30 also
is anticipated  (Howarth,  C., personal communication).   At  publication,
FCC  had just  approved GEOSTAR  Corporation's  application for use of the
requested  frequencies.  Candidate users are urged to confirm FCC  approval,
verify  the latest satellite/ground  station operating schedule, and obtain
further information on transceiver availability.

Navstar GPS—

     The Navstar  Global  Positioning System (GPS)  is  a  second-generation
satellite navigation system currently under development by  the U.S. Department
of Defense.  Its purpose  is to provide precise,  continuous, worldwide,
all-weather, three-dimensional  navigation  for land, sea, and  air  applications.
Under current plans, 18  satellites will be  launched into three co-planar
orbits 120° apart  to provide  continuous  transmission of time, three-dimensional
position, and velocity  messages to system users.  The GPS  satellites  transmit
at 1,227.6 MHz and 1,575.4 MHz to permit the measurement and  correction
of ionospheric refraction error.   Five developmental satellites  currently
are  in orbit,  providing approximately 4 h of coverage twice daily, separated
by a  12-h period.   Continuous two-dimensional positioning information should
be available with 12 satellites by February, 1988, and  three-dimensional
coverage is projected  for  late 1988 or early  1989 (DeGroot,  L.,  personal
communication; Stansell  1984).  The system  consists of the satellites  in
12-h, 20,200-km (12,552-mi)  orbits, a U.S. master control  station,  several
monitoring stations, and  small,  lightweight, relatively inexpensive receivers.
Signals received from  any  four Navstar  satellites are demodulated, time-
correlated, and processed  to obtain precise position  information.

      Two  levels of positioning accuracy are  achievable with the Navstar
GPS  system.  The lower level  is  obtained from the  Standard Position Service
(SPS) using  the  coarse  acquisition or "C/A code."   When the system becomes
fully operational, navigational accuracy from these  signals should be  approxi-


                                   C-39

-------
                                                       Station Positioning
                                          Appendix C - Positioning  Methods
                                                               August  1986
mately  100  m (328 ft) two-dimensional rms,  or a circular probable  error
(CEP) of 40 m (131 ft)  (Montgomery,  B.,  personal communication).  More
accuracy  can be achieved  using the Precise  Positioning Service (PPS) or
"P-code" (i.e.,  8-9 m two-dimensional  CEP).  Additional positioning accuracy
can  be  achieved by operating in a differential  mode, in which receivers
on a vessel  and at a known onshore location  simultaneously receive the
satellite signals.  The onshore receiver  is  calibrated.  Bias corrections
based on signals received at the fixed  station  are  transmitted to the mobile
receiver.  These area-specific corrections yield more accurate positional
determinations.  With differential GPS, a two-dimensional  position should
be definable within a range of 2-5 m (6.6-16.4  ft)  (Montgomery 1984;  Stansell
1984).

     Due  to  the present lack of  full-time coverage, both SPS and PPS are
available to military and civilian users.  However, the  government  intends
to encrypt the P-codes, allowing use only by the  military and other authorized
users [e.g., National Ocean Industries  Association (NOIA)  members].  NOIA
has  proposed that all P-code receivers  be owned and operated by a single
service company that would  provide  both the equipment  and personnel when
national  security conditions are met and  no  other reasonable navigational
alternatives are available (Stansell 1984).   Under U.S. Department of Defense
policy, access to PPS is available to civilian  users operating in the "national
interest" (Anonymous 1985).  Equipment, cryptography,  and  support services
will  be provided under contract to the user.   However,  it would appear that
general users in coastal areas will be limited to  C/A code equipment.   It
may  therefore  be necessary to operate in  the  differential mode in order  to
achieve  the positional accuracies needed for survey and monitoring requirements.

     Because the GPS system  is in a developmental  stage, cost estimates
for the  equipment are difficult to make.  Several  major equipment manufacturers
are  in  the process of  designing  receivers with  varying  capabilities,  and
a limited number of models  now are available.   Some manufacturers envision
that receivers with 100-m (328-ft) accuracy  will cost  less than $500 when
mass produced (e.g.,  for automobiles).  At  the  other extreme, for $140,000
Texas Instruments sells  the TI4100  Navstar  Navigator, said to be capable
of slow dynamic positioning within a  few  meters, speed within  tenths  of
a knot, and  time to the  usec  (Montgomery 1984; St.  Pierre, R.,  personal
communication).  Motorola anticipates that  the  initial  cost of two  stations
needed to operate in  the  differential mode  will be in the $100,000-range
(Sheard,  S., personal communication).   Magnavox has  a five-channel  T-Set
GPS  Navigator, with real time  differential  GPS  operation as a planned option.
A two-unit system, excluding  communications link, costs  approximately  $100,000
(Driscoll, C.,  personal  communication).   Rockwell  International sells a
prototype C/A code receiver for  $17,500  and anticipates that GPS  receivers
will cost less  than $10,000  by 1988  (DeGroot, L., personal  communication).
Tracer  expects  initial  models to sell  for less than $10,000,  falling  to
around  $1,000 in 3-5  yr  (Murphy,  W., personal  communication).
                                   C-40

-------
                                                       Station Positioning
                                          Appendix C -  Positioning Methods
                                                              August 1986


     Using the C/A code in a differential mode, the Navstar GPS positioning
system should provide  accuracies within +3 m (+9.8 ft).  Errors of  1  m
(3 ft) at  a  range of ZOO km (12.4 mi)  from the reference  station are predicted
for daytime  operation.  Smaller errors are expected at night (Kalafus 1985).
However,  testing with the actual  satellite signals modified to reduce achievable
accuracy  will  be required  to confirm this capability.   The  two  required
stations  may be expensive at first, but competition and  the large potential
market are expected to drive the price much lower.

SERIES—

     The  SERIES positioning and navigation system (Satellite Emission Range
Inferred  Earth Surveying) has evolved  from a technique originally developed
to measure movements in the earth's crust for tectonic studies  and earthquake
predictions.   The system enables use of signals from existing Navstar  GPS
satellites  with no knowledge of either the P or C/A codes by simultaneous
pseudoranging to multiple satellites in a differential  mode.  Accuracies
on the order of 2 m (6.6  ft) CEP  or 5 m (16  ft)  two-dimensional rms are
said to be achievable for vessels in a slow dynamic mode (0-10 kn)  (ISTAC
1984). This is  accomplished  by comparing Navstar signal arrivals at one
receiver to the same signals received at  another shore-based geodetic  reference
mark.  By explicit differencing of  the frequency and phase measurements
observed  from each satellite, it is  apparently possible to eliminate  the
effects  of  Navstar signal  variations that secure the satellites from an
unauthorized user in an autonomous,  single-receiver, real time  mode.  SERIES
receivers are codeless spectral compressors  that simultaneously receive
and incorporate  spread-spectra modulations of  several Navstar satellites
and then  extract the  frequency and  phase of each satellite in view.  Data
are processed by a combination of Doppler positioning and phase ranging.

     System  components include  an unattended,  fixed  reference station,
a marine  vessel  station, and a reference  link (via satellite or terrestrial
radio) to transfer the reference  station signal to the  marine vessel for
real time positioning.   The reference  station includes an  SERIES GPS receiver
(MPS-1),  a  rubidium frequency reference, and  a customer-supplied radio
transmitter  (200 Hz analog signal).  The marine station consists  of the
GPS receiver with an  integrated Navstar C/A-code-correlating  subsystem
for broadcast satellite ephemerides, a rubidium frequency  reference,  a
microprocessor  for real time position computations, and  a  customer-supplied
receiver  for  fixed reference station signal  reception.  A  standardized
RS-432 interface is provided for communication of the position information
to the vessel's  navigation system.

     The  initial  MPS-1 units  are estimated to  cost $205,000 for a  ship
system and $82,000 for the fixed base  system required  for  differential
mode  operation (Whitcomb,  J., personal communication).  The system  will
probably  be  available on a lease or  service-contract basis, as  well.
                                  C-41

-------
                                                      Station Positioning
                                          Appendix  C - Positioning Methods
                                                              August  1986


     When  operational, the SERIES  system may offer  a  high-accuracy alternative
to the Precise Positioning Service of Navstar GPS  or to differential mode
operation with  the C/A code.  Although  the projected cost for an initial
unit is high  (to recover research  and development expenditures), cost should
decrease  with  further production.   Potential users should confirm price,
verify accuracies achievable  from  a mobile marine vessel, and  identify
specific data-link components that need-to be supplied by the user.

Aero Service  GPS—

     The Aero Service Division  of Western  Geophysical Company of America
is developing a codeless satellite navigation system  for marine positioning.
The  system will  enable real  time  navigation in the slow dynamic mode to
less than  5 m (16.4 ft).  For vessels  in the semi-static mode (on  station)
for  30  min,  the system will  reportedly yield a position  fix accuracy of
less than  1 m (3.3  ft) (Mateker,  E.,  personal communication).  The  system
will  require one unit onshore  at  a fixed position, and one unit on the
survey vessel.  Continuous real time  communication  between the two  systems
may  or  may  not  be necessary, depending on steps  by the U.S. Department
of Defense  to  degrade certain elements of  the GPS satellite signals.  Prototype
testing was  scheduled for September,  1985, with  production  systems planned
for early  1986.

     This future  system is worthy of note because  it will  probably be based
on methods now used for land surveying in the MACROMETER Interferometric
Survey System.   The MACROMETER  is  designed for very precise positioning
in a fixed-point  (static) mode, using  GPS signals but  no codes.   Company
tracking  stations (Phoenix, Arizona and Woburn,  Massachusetts) will follow
up to six  satellites  simultaneously.   The MACROMETER  is capable of  providing
position  in  terms of latitude,  longitude, and ellipsoidal  height  to within
a few parts  per  million  of the distance  from  a  reference  point.  In the
continental  U.S., 1-3  h  of observation will yield  first-order (1:100,000)
accuracy in  all  three coordinates using  the GPS  satellites.  Data for a
second-order or third-order survey  point  may be  collected in  less  than
1 h  (e.g., as little  as  15 min of observation).

     The  MACROMETER  surveying  system includes two or more field  units and
an office data processor.  Each V-1000  field unit includes  an  antenna,
receiver, tape drive,  30 m of cable, time  receiver, power package,  battery
chargers,  generator,  and DC/AC inverter.   The P-1000 Office  Data  Processor
computer  subsystem  includes hard and floppy disk drives, video  terminal,
tape drive,  printer,  modem, and  all  necessary software.   A  1,000 series
two-unit  surveying  package is  currently  priced at  $235,000.   The system
can  be  rented for $5,000 per V-1000  unit  per month,  with  a  2-mo minimum
on each unit.
                                   C-42

-------
                                                       Station Positioning
                                          Appendix C - Positioning  Methods
                                                               August  1986
RANGE-AZIMUTH  SYSTEMS
     A number of  hybrid positioning  systems combine positional  data from
various sources to obtain fixes.  Such systems  usually involve the intersection
of a visual  1 ine-of-position with an electronic line-of-position (Umbach
1976).   Visual  data may be in the  form  of sextant angles  or theodolite
azimuths.   Electronic  positional data 'are  normally obtained from a microwave
system.

     Of particular interest  for survey and monitoring programs are dynamic
positioning systems that require only a single shore station and  that  use
the  simultaneous  measurement of angle  from a known direction and range
to the  survey vessel. This range-azimuth method has the advantage of circular
coverage  around the shore  station (Figure C-8).  A single station minimizes
logistical requirements  and  geometric limitations.  Line-of-position  inter-
sections  are the  ideal 90° everywhere  within the coverage area.  Growth
in the error ellipse is  due  only to distance from the shore station because
of its independence  of the absolute  azimuth angle.  Accuracy improves as
range decreases, even  fairly close to the shore station, and only one  unob-
structed line-of-sight is needed.  However,  such systems allow only a single
user.

     Characteristics of three representative systems are  summarized in
Table  C-12.  Two have  fully automatic shore stations,  requiring attendance
only during setup and  alignment.  The  third requires an onshore operator
at all times.  For multiple-day surveys,  an automatic  station eliminates
the  tedium of  continuously  tracking the survey vessel.  The  systems offer
much greater flexibility than range-range or hyperbolic systems with multiple
onshore stations.   Where positioning requirements extend  into ports, estuaries,
or up rivers, the single-station systems offer distinct  advantages in covering
restricted or  congested survey  areas and in establishing  an unobstructed
signal path.  Each system  is  distinctive  in  either  its  operating  medium
(optical, microwave,  laser)  and/or  procedure (i.e., manual or  automatic
tracking).  The added  costs of these systems over standard microwave  systems
may  be justified where the system  also can be used  for onshore work  during
nonsurvey  periods.

AZTRAC

     The ODOM Offshore  Survey  AZTRAC is a semi automated optical angle-measuring
and  transmitting  system which  can  be used in conjunction with an  independent
distance-measuring system to  position  a vessel.  The AZTRAC  system consists
of a modified Wild T16  theodolite,  an onshore transmitter,  and a  vessel
receiver   The  theodolite has  an infinite tangent drive and  provides information
in a digital format.  Typically,  the separate distance-measuring  system
consists  of a  microwave master  receiver  on the  vessel  and a  remote  unit
transmitter  located at  the theodolite.
                                    C-43

-------
Figure C-8.   Range-azimuth positioning system area of coverage.
                           C-44

-------
                                 TABLE C-12. RANGE-AZIMUTH POSITIONING SYSTEM CHARACTERISTICS
SYSTEMS
AZTRAC
Odom Offshore Surveys
Baton Rouge, Louisiana
(504)769-3051
POLARFIX
Krupp Atlas-Elektronik
Webster, Texas
(713)338-6631
ARTEMIS
Andrews Hydrographies
FREQUENCY RANGE
Optical Visible
and
microwave
(typical)
Laser 5000 m
(904 ran)
UHF Telemetry
(406-470 MHz)
9.2-9.3 GHz 10-1400 m or
200-30,000 m
NOMINAL ACCURACY
AZIMUTH RANGE POSITION
0.01 Both depend on accuracy of
ranging system used
0.01° i0.1m±0.1m/km ±0.1m±0.2m/km
0.03° 0.5 m short Not stated
1.5 m long
COST
$22,500 plus
ranging
system cost
cost
$100,000
$70,000 to
$75,000
Houston, Texas
(713)558-2236

-------
                                                      Station Positioning
                                          Appendix C - Positioning Methods
                                                              August 1986
     For a survey, the  AZTRAC theodolite and  transmitter are set up at
a known  position and the theodolite  is  zeroed on a known  azimuth line of
backsight.  The  theodolite operator sights  and tracks the survey vessel's
ranging  antenna or transponder.  As the vessel  moves, the microwave ranging
system continuously  measures distance  between the shore station and the
vessel.   The tracking  motion produces pulses that are decoded by the AZTRAC
transmitter and  displayed to the  theodolite operator as the angle to the
vessel  from the reference azimuth  line.   The angle also  is  converted to
BCD serial format and is used to  activate  the transmitter,  which sends
the information to the survey  vessel.   The AZTRAC receiver  converts the
angle information to parallel format and displays  it for manual recording.
It simultaneously outputs the angle as serial  data  for automated recording
or processing by any available  on-board computer  or  plotting system.

     AZTRAC is designed to operate with  most  microwave ranging systems.
Two AZTRAC units working in  an  azimuth-azimuth mode  can provide positioning
in survey areas where reflections from metal  structures or electrical noise
from radar and  other transmitters  limit use of microwave ranging.  The
Wild T16  theodolite  has a 30X magnification,  27-m (89-ft)  field of view
at 1,000  m (3,280 ft), and an angular resolution  of 0.01° (36 arc sec); at
a distance of 5 km (3.1 mi), this  corresponds to  an absolute arc length
error of 0.9 m (3.0 ft).

     The National Ocean Survey recently examined  range-azimuth positioning
of a vessel moving at a  nominal  speed of 6 kn (11.1  km/h)  at ranges of
up to 3,000 m  (9,842 ft).   Instruments included a  Wild T2 theodolite,  from
which angles were manually recorded onshore, and the AZTRAC,  whose angles
were recorded  on the vessel.  Pointing errors (68 percent  probability)
of these two instruments  were  found to be approximately  1.3 m (4.3 ft),
independent of  range when  standard  deviations of right and left movement
data were pooled  (Waltz 1984).

     The theodolite  was sited on a white Del Nprte Trisponder transponder
on the moving vessel.  Visibility  was good during the  2-day survey (off
Monterey,  CA),  with calm mornings giving way to afternoon winds of 15 kn
and 0.6-0.9 m (2-3 ft) seas.  Range  could have been extended  much  farther
under  such conditions,  particularly  if color had been added to the vessel
target (Waltz, D.A.,  personal communication).   U.S.  Army Corps of  Engineer
users  confirmed the effectiveness and reliability of  the  AZTRAC system
in conjunction with several  different microwave  ranging systems (Ard, R.,
personal  communication).   The system also has proven  effective in  port
and harbor surveys when it was  impossible to achieve optimal shore  station
geometry for range-range or hyperbolic  operation.

     The current  cost of the AZTRAC  alone is $22,500  (Apsey,  B., personal
communication).   A system consisting of the AZTRAC, a  Motorola Falcon 484
with one reference  station  (approximately $32,000),  and  an interface unit
($10,000) would cost $64,500.   Thus,  provided the required  range  is achievable
                                  C-46

-------
                                                       Station Positioning
                                          Appendix C - Positioning Methods
                                                              August 1986
under anticipated visibility  conditions, the additional  versatility of
the AZTRAC  can  be  realized  for  $25,000 above  the  $39,300 cost of  a  two-
station  Falcon  484 range-range microwave system.

POLARFIX

     POLARFIX  is  a dynamic range-azimuth positioning system  by Krupp Atlas-
Elektronik.   The  system uses a scanning (30° horizontal)  pulsed laser beam
from a single,  fixed, onshore tracking  station to follow the survey  vessel
(up to  10°/sec) and to transmit range and angle information via telemetry
link.   The  system  incorporates a fully  automated onshore tracking  station,
which requires  no  attendance beyond initial station setup and azimuth refer-
encing.   The  shore station can locate  the  mast-mounted prism reflectors,
follow  the vessel,  and,  if  necessary, relocate  the vessel by performing
a routine search pattern based on a record  of tracking history.  The shore
tracking  station  consists of a  laser-sensing head mounted  on a conventional
survey tripod,  linked by cable to an integrated  control unit that  houses
data control, transmission, and  telemetry transceiver equipment.  An integral
control  unit  (including a display and a second telemetry receiver), keyboard
terminal,  printer,  telemetry  antenna, and prism reflector assembly are
onboard the vessel.

     Under clear operating conditions,  a 3-km (1.9-mi) range  using a  Class  I
laser or a 5-km (3.1-mi) range using a Class  Ilia laser may be  selected.
In  foggy weather, range  is  said to be 1.5 times visible  range, due to use
of the pulsed infrared  laser.  Maximum range  achievable varies  with the
prism assembly used to reflect  the tracking station's laser  beam.  Single-,
dual-, and triple-ring omnidirectional prism assemblies can  be stacked
on the vessel antenna.  For average weather conditions, a two-prism assembly
(each with five reflectors) gives approximately 3.5 km  (2.2 mi)  of  range.
This  can be  extended to  approximately 5.0 km  (3.1 mi)  with  the addition
of more assemblies.  Although  a  5  km (3.1 mi)  distance from  shore will
cover most of the  Sound, this does add  limitations to range in  some areas.

     Range accuracy is reported as 0.1  m +0.1 m/km (0.3  ft +0.5 ft/mi)
of  measured  range.  Azimuth accuracy is said to be 0.01°  or  better.  The
resulting  positional accuracy at  1,  3, and  5 km (0.6,  1.9,  and 3.1 mi)
is  approximately 0.3, 0.6,  and  1.0  m (1, 2,  3.2 ft), respectively.  The
positional algorithm  given  is +0.1 m +0.2 m/km (+0.3  ft +1.1 ft/mi).   Current
cost of the system is $100,000  (Guillory, J., personal communication).

ARTEMIS

     The ARTEMIS by  the Christiaan Huygenslabortorium  (Holland) is a distance-
bearing type  of microwave positioning system  capable  of measurements at
ranges  of 10 m (32.8  ft) to 30  km (18.6 mi),  and angles  from 0°  to 360°
from a single  fixed shore  station.  Accuracies at  the two-sigma or 95  percent
level  are given as  +1.5 m (4.9 ft) distance, and  +0.03° azimuth, equivalent
to +0.5 m/km  (2.6  ft/mi).

                                   C-47

-------
                                                      Station Positioning
                                          Appendix  C - Positioning Methods
                                                              August 1986
     Angle measurements  are based on automatic tracking antennas on the
vessel  and at the shore station.  Once locked, the two  antennae move  always
pointing  towards  each other.  A maximum combined tracking speed of 3°/sec
is allowable to achieve the specified angle  error.  The direction  of the
fixed  station  antenna is accurately measured with a precision shaft coder,
which is mechanically coupled to  the ma-in shaft  of the antenna.  Measured
angle data are transferred to the mobile station via  the established continuous
microwave channel.  The same microwave link  is  used  to measure distance
by controlled interruption of the microwave signal.  Both angle and distance
usually are displayed on  the Mobile Control  Data Unit, although  readout
at the  shore  station also  is  feasible.   The microwave link also  is used
for voice communication between  the two stations without disturbing the
data being transmitted.

     The  vessel's  positioning equipment  consists of  the Mobile  Control
Data  Unit (MCDU), a Mobile Antenna Unit  (MAU), the antenna, cables,  a telephone
handset,  and  a speaker.  The  shore  station  consists  of a Fix Antenna Unit
(FAU),  Fix Control Data Unit  (FCDU),  an antenna, cables, a telephone handset,
a  speaker, and a telescope for initial  directional  alignment.  The  shore
station  requires  attendance only for setup, referencing, and periodic  battery
checks.  The  bearing is  electronically referenced  to a geodetic  grid by
siting the unit  in  a  known reference direction and manually  adjusting the
observed direction  readout to the correct value.

     Limiting  factors (common to all  microwave systems) include  radio  line-
of-site conditions  (obstacle-free for optimum performance)  and multipath
interference due to  reflections from the  sea  surface.  The latter can be
reduced by proper  adjustment of antenna  heights.  Where survey area  traffic
is heavy, signal  interruptions could unlock  the two tracking antennae.
Manual  relocking by a shore  station attendant or  automatic relocking using
autosearch,  an option available at added  cost, would then be necessary.
Although  signals generally are unaffected  by rain  and  fog,  there  is some
range  reduction during heavy rains or snowfalls.

     The  present system  costs  $70,000-$75,000,  depending on the options
selected  (Coupe, C.,  personal communication).  The  system  may be rented
from Andrews Hydrographies, Inc. (Houston, Texas)  for approximately $850
per  day.
                                   C-48

-------
                                                       Station Positioning
                                          Appendix C -  Positioning Methods
                                                              August 1986
                               APPENDIX C

                               REFERENCES


Anonymous.   1981.   Satellite  program surveys 10  sites  simultaneously.
Sea Technology.  22(5):41.

Anonymous.   1985.   Precision  code  unavailable to civil  sector GPS users.
Sea Technology.  26(1):77.

Apsey,  B.   10  January 1985.   Personal  Communication (phone  by Dr. William
P. Muellenhoff).  Odom Offshore Surveys,  Inc., Baton Rouge,  LA.

Ard, R.   18 January 1985.  Personal  Communication (phone  by Dr. William
P. Muellenhoff).  U.S. Army Corps of Engineers  San Francisco District,
San Francisco,  CA.

Baker,  W.F.  11 January 1985.   Personal  Communication (phone  by Dr. William
P. Muellenhoff).  Teludist, Inc.,  Mastic  Beach, NY.

Blizard,  M.M.,  and O.C. Slagle.  1985.   LORAN-C West Coast stability study.
Technical  paper presented at the Fourteenth  Annual  Wild Goose Association
Convention,  October 23, 1985, Santa Barbara, CA.

Booda, L.L.  1984.   Civil  use of Navstar  GPS, a matter  of debate.   Sea
Technology.   25(3):17-18.

Bowditch,  N. 1984.  American practical  navigator.  An epitome  of navigation.
Defense Mapping Agency Hydrographic Center.

Bruckner, D.C.  1985.  Differential  Loran-C:  estimator improvement and
local system implementation.  Technical  paper  presented at the Fourteenth
Annual  Wild Goose Association Convention,  October 23, 1985,  Santa Barbara,
CA.

Buchanan,  C.  31 December 1984.  Personal Communication (phone by Dr. William
P. Muellenhoff). Del  Norte Technology,  Inc., Euless,  TX.

Canadian  Coast Guard.   1981.  A primer on  Loran-C.  TP-2659.   Aids and
Waterways,  Canadian  Coast Guard, Department of Transportation, Ottawa,
Ontario, Canada.  29  pp.

Clark,  D.  1951.  Plane  and  geodetic  surveying  for engineers.   Constable
and Company, Ltd.,  London, UK.  636 pp.
                                   C-49

-------
                                                        Station Positioning
                                           Appendix  C  -  Positioning Methods
                                                               August 1986


Coupe,  C.J.   23 January  1985.  Personal Communication  (phone by Dr. William
P. Muellenhoff).  Andrews Hydrographic Services Inc., Houston, TX.

Davis, R.E.,  F.S.  Foote,  and J.W. Kelly.  1966.  Surveying theory and practice.
McGraw-Hill  Book Co.,  New York, NY.  1096 pp.

DeGroot,  L.   15 January  1985.  Personal Communication  (phone by Dr. William
P. Muellenhoff).  Rockwell  International, Cedar Rapids, IA.

Doughty,  R.A., and W.K. May.  1985.   Differential  Loran-C...where do we
really need it.  Technical  paper presented at  the  Fourteenth Annual  Wild
Goose Association Convention,  October 23, 1985, Santa Barbara, CA.

Driscoll,  C.   15 January  1985.  Personal Communication  (phone by Dr. William
P. Muellenhoff).  Magnavox, Torrance, CA.

Dungan, R.G.   1979.   How  to get the most out of Loran-C.   SG 54.   Extension
Marine Advisory Program,  Oregon State University, Corvallis, OR.   12 pp.

Folk, L.  21  March  1985.  Personal  Communication (phone by Dr.  William
P. Muellenhoff).  Kuker-Rankin, Inc., Seattle, WA.

Gazley, L.  17 December 1985.  Personal Communication (phone by Dr.  William
P. Muellenhoff).  U.S.  Coast Guard  Office of Navigation, Radionavigation
Division,  Washington,  DC.

Guillory,  J.  28 January  1985.  Personal Communication  (phone by Dr. William
P. Muellenhoff).  Krupp Atlas-Elektronik, Webster, TX.

Harris, E.  3  January 1985.  Personal Communication  (phone by Dr. William
P. Muellenhoff).  Del  Norte Technology  Inc., Camarillo, CA.

Hempel , C.   10 January 1985. Personal Communication  (phone by Dr. William
P. Muellenhoff).  Cubic Western Data,  San Diego, CA.

Hoeber, J.L.   1981.   An update on  worldwide  navigation  systems  -  the present
and  the year 2000.  Sea Technology.  22(3):10-13.

Howarth,  C.   16 January 1985. Personal Communication (phone by  Dr. William
P. Muellenhoff).  Geostar Corp., Princeton,  NJ.

Ingham, A.E.   1975.   Sea surveying.  John Wiley and Sons,  New York,  NY.
306  pp.

ISTAC.  1984.  Marine Positioning  Sensor.   MTS-1.   Technical  Data Sheet
ISTAC-SERIES.   ISTAC,  Inc., Pasadena,  CA.   p.  47.

Kalafus,  R.M.   1985.  Differential  GPS standards.   Satellite navigation
system  is a promising  new tool of  high accuracy.  Sea  Technology 26:52-54.

                                   C-50

-------
                                                       Station Positioning
                                           Appendix  C - Positioning Methods
                                                               August 1986
Jolly,  J.   31 December  1984.  Personal  Communication  (phone by Dr. William
P. Muellenhoff).   Jon  B.  Jolly, Inc., Seattle,  WA.

Jolly,  J.   15 January 1985.   Personal  Communication  (phone by Dr. William
P. Muellenhoff).   Jon  B.  Jolly, Inc., Seattle,  WA.

Maloney,  E.S.  1978.  Duttons navigation  and piloting.  Naval  Institute
Press, Annapolis,  MD.  910 pp.

Mateker,  E.J., Jr.   17  January 1985.   Personal  Communication  (phone by
Dr. William P. Muellenhoff).  Litton Aero Service,  Houston, TX.

McDonnell, Jr.  1983.  Total station survey.   Point of  Beginning.   8:16-30.

Moffitt, F.H.,  and H.  Bouchard.   1982.  Surveying.  Harper  and Row Publishing,
New York,  NY.   834 pp.

Montgomery, B.O.   1984.  Navstar GPS  - A  giant step  for  navigation and
positioning.  Sea Technology.   25(3):22-23.

Montgomery, B.  15 January  1985.  Personal Communication  (phone by Dr. William
P. Muellenhoff).   Texas  Instruments, Lewisville, TX.

Moyer,  C.  15 January  1985.   Personal Communication  (phone  by Dr. William
P. Muellenhoff).   Motorola,  Inc., Tempe, AZ.

Murphy, W.  15 January  1985.  Personal Communication  (phone  by Dr. William
P. Muellenhoff).   Tracer, Inc., Austin, TX.

Newcomer,  K.E.  21 January  1985.  Personal Communication  (phone by Dr. William
P. Muellenhoff).   Navigation Sciences, Inc.,  Bethesda,  MD.

Panshin,  D.A.  1979.  What  you should know about Loran-C receivers.   SG 50.
Extension Marine  Advisory  Program, Oregon State University, Corvallis,
OR.  8  pp.

Ryan,  R.   11 January 1985.  Personal  Communication  (phone  by Dr. William
P.  Muellenhoff).  USCG  Pacific Area, Government Island,  Alameda, CA.

Sheard, S.   15 January  1985.  Personal  Communication  (phone  by Dr. William
P. Muellenhoff).   Motorola  Inc.,  Tempe, AZ.

Slagle, 0.   31 October  1985.  Personal  Communication  (phone  by Dr. William
P. Muellenhoff).   U.S. Coast Guard  Research  and Development  Center, Avery
Point,  Groton, CT.
                                    C-51

-------
                                                        Station Positioning
                                           Appendix C - Positioning Methods
                                                                August 1986


Stansell,  Jr., T.A.   1984.   GPS marine user equipment.  Magnavox Technical
Paper MX-TM-3381-84.   Presented  at  the Global  Civil Satellite Navigation
Systems Conference,  The Royal  Institute of Navigation, London, UK.

St.  Pierre,  R.   15 January 1985.   Personal Communication (phone by Dr. William
Muellenhoff).   Texas  Instruments, Lewisville, TX.

Sturgill, D.   18  November 1985.   Personal Communication  (phone by  Jeff
H. Stern).   Municipality of Metropolitan Seattle, Seattle, WA.

Tetra Tech.  1986.  Evaluation of survey positioning methods  for nearshore
marine  and estuarine waters.   U.S. EPA Office of Marine and Estuarine Pro-
tection, Washington,  DC.  Tetra  Tech,  Inc., Bellevue, WA.  107 pp.

Umbach,  M.J.   1976.   Hydrographic manual, fourth edition.  U.S.  Department
of Commerce, NOAA, Rockville,  MD.  414 pp.

U.S. Coast  Guard.   1974.  Loran-C user handbook.  CG-462.  U.S. Coast Guard,
Washington, DC.  25 pp.

U.S.  Department  of  Defense  and U.S. Department of Transportation.   1984.
Federal radionavigation plan.   DOD-4650.4, DOT-TSC-RSPA-84-8.  Washington, DC.

Waltz, D.A.  1984.  An error  analysis  of range-azimuth positioning,  pp. 93-
101.  In:  Proc. of National  Ocean Service Hydrographic  Conference.   April
25-27, Rockville,  MD.

Waltz,  D.A.  18 January 1985.  Personal Communication (phone by Dr. William
P. Muellenhoff).  National Ocean Service, Norfolk, VA.

Whalen,  W.L.   1984.   Geostar positioning system using satellite technology.
Sea Technology.  25(3):31-34.

Whitcomb, J.H.  14 January 1985. Personal Communication (phone by Dr.  William
P. Muellenhoff).  ISTAC, Inc., Pasadena, CA.
                                    C-52

-------
           APPENDIX D





EVALUATION OF POSITIONING METHODS

-------
                                                       Station Positioning
                                           Appendix  D  - Methods Evaluation
                                                              August 1986
                               APPENDIX D

                    EVALUATION OF  POSITIONING METHODS
     Methods  to position at stations  in Puget Sound for collection of environ-
mental  samples are addressed in this  appendix.   After an overview  of  each
method  is  presented, the  available positioning methods  are evaluated and
comparative information is presented to facilitate selection of a method
appropriate to meet sampling objectives.  Positioning methods and associated
equipment are described in detail  in  Appendix C.

POSITIONING PROCEDURES IN USE IN PUGET SOUND

     Equipment,  procedures, advantages,  and limitations of  positioning
methods used  in Puget Sound were identified  from interviews with  various
governmental and private groups that conduct environmental sampling within
Puget Sound  (Table D-l).   These methods, their accuracies as expressed
by the  field personnel who work with them, and the limiting restrictions
are summarized in Table D-2.  The majority of  the sampling done  in Puget
Sound in recent years has been accomplished with the following five positioning
methods:

     •    Sextant angle resection

     •    Variable range radar (VRR)

     •    Loran-C

     •    Combinations of VRR, visual fixes,  and Loran-C

     •    Microwave systems.

     Many  smaller  research  vessels in  Puget Sound  are  equipped  with VRR
and positioning by this method is usually used when VRR fixes are considered
adequate for station location definition.  However, if  higher accuracies
are required, VRR fixes  are supplemented by visual  line-of-site fixes,
bottom  depth readout, and Loran-C.  Loran-C  cannot initially be used to
locate a station  in most  areas of Puget Sound.  Loran-C coordinates are
recorded,  however, to help improve repositioning accuracy.  Overall, the
combination of methods should increase accuracies closer  to shore  (visual
fixes) and with steeper bottom slopes (fathometer).

     Of  the  other  methods  practiced  in Puget  Sound, sextant angles are
used the least.   Because of  visibility effects,  it  is difficult  to  estimate
the accuracy  of this method, which can change considerably between  stations.

                                   D-l

-------
          TABLE  D-l.   INTERVIEWED  GOVERNMENTAL  AND  PRIVATE  GROUPS
                    THAT CONDUCT SAMPLING  IN  PUGET  SOUND
GOVERNMENTAL AGENCIES AND INSTITUTIONS

Federal

National Marine Fisheries Service (NOAA)
National Ocean Survey (NOAA)
Pacific Marine Environmental Laboratory (NOAA)
U.S. Army Corps of Engineers
U.S. Environmental Protection Agency

State

Department of Ecology
Department of Natural Resources
Department of Social and Health Services
University of Washington, College of Ocean and Fisheries Sciences

Local

Municipality of Metropolitan Seattle (Metro)


PRIVATE FIRMS

Battelle Northwest Laboratories
Cooper Consultants,  Inc.
Evans Hamilton, Inc.
EVS Consultants
Northern Technical Services, Inc.
Parametrix,  Inc.
URS Engineers


RESEARCH VESSELS

Clifford A.  Barnes (UW)
DNR III (WDNR)
Kittiwake (Private)
Liberty (Metro)
McArthur (NOAA)
Harold K. Streeter  (NOAA)
                                      D-2

-------
                                        TABLE  D-2.   CHARACTERISTICS OF POSITIONING METHODS
                                                 USED FOR SAMPLING  IN  PUGET  SOUNDa
o
co
Method
Sextant Angle
Resection

Variable Range
Radar (VRR)
Loran-C
VRR Visual Fixes,
Loran-C

Microwave Systems
Areas
Open Areas
Nearshore
All
Limited
Open Areas
Nearshore
All
User Reported Accuracy Ranges (+m)°
Absolute Repeatable Restrictions Use Among Local Groups
100
10
70
40
30
10
2.5
- 200
- 20
- 120
- 100
- 100
- 40
- 10
<100 Visibility, landmarks
Similar <2 - 3 km from shore
20 - 50 Increases with distance,
featureless topography
10 - 30 Only in Loran-C
mapped areas
20 Repeatability drops in areas
with no Loran reception
Similar <3-4 km from shore, Loran not as
critical to repeatability
1-3 Transponder locations
Low
Low
Medium
Medium
High
Medium
High
             a As identified  through user interviews.

             b Reported accuracy ranges are consistent  with manufacturer claims with
             the exception that sextant accuracies only achieve highest accuracies near-
             shore.

-------
                                                       Station  Positioning
                                           Appendix D - Methods Evaluation
                                                               August 1986


Therefore,  sextants are used mainly during region characterization  studies
when location  is  not critical.  When  strict absolute accuracy  is needed,
microwave  systems are  frequently used.  In areas that have  been mapped
for Loran-C range lines,  absolute accuracies  as good as or  better than
VRR can  be achieved with  Loran-C.   As  more stations are occupied in these
mapped areas,  comparisons of Loran-C  signals with other fixes  help define
areas of  signal  warpage  from land  ar  building interactions,  increasing
absolute accuracy.  These systems and their methods are discussed in detail
in Appendix C.

CANDIDATE SYSTEM  OVERVIEW

     Methods  other  than those  commonly used in Puget Sound for  station
positioning are available.  Characteristics,  major advantages, and major
disadvantages of each  method listed below are  summarized in this section.
Detailed descriptions are presented1in Appendix C.

     0    Multiple horizontal angles

          Theodolite
          Sextant angle  resection

     •    Multiple electronic ranges

          Variable range radar
          Distance-measuring instruments
          Microwave  systems
          Loran-C
          Satellite  ranging

     t    Range and  angle

          Theodolite and EDMI
          Total station
          Range-azimuth  positioning systems.

Multiple Horizontal  Angles

     Theodolites have  the  necessary angular acuracies at the anticipated
maximum ranges.   They are commonly used as surveying  instruments and cost
$2,000 (30-sec accuarcy) to  $4,000 (10-sec accuracy).  At  least  two theodolites,
two operators, a  siting  target on the vessel,  and a  three-way communications
link to coordinate  fixes are  required (see Appendix C, Figure C-l).  Theodolites
can be used only  during  daylight hours of good visibility.

     Sextant  angle resection  techniques offer adequate angular accuracy
( +  10 sec) and sextants  costs $1,000-$5,000.   A three-arm  protractor  is
required for plotting  positions.   Two oprators  should take simultaneous
fixes on moving vessels.  Because  the operators  are  on board, a separate


                                    D-4

-------
                                                       Station Positioning
                                           Appendix D - Methods Evaluation
                                                              August 1986


communication  link is not necessary and  they can also serve as crew.  However,
the method  requires highly visible shore targets  and is therefore useful
only during daylight hours of good visibility.  In addition,  it  is difficult
for even an experienced  operator to shoot an  accurate  fix  from  a moving
platform in adverse weather.

Multiple Electronic Ranges

     Positioning with  multiple electronic VRR ranges will  be less  accurate
over anticipated distances than electronic positioning methods.   Equipment
costs  range from $4,000  to $10,000.  Weather and visibility  rarely  limit
use and extra  personnel are not needed  to help navigate.

     Positioning with  Electronic  Distance-Measuring  Instruments (EDMI)
offers adequate accuracy but marginal range  beyond 3 km (1.9  mi)  without
multiple  prisms.   EDMI systems cost $8,000-$15,000 apiece for  long-range
units and approach $40,000 for  systems with  prisms.   EDMIs  require two
staffed stations, a three-way communications  link to coordinate fixes,
and multiple prism assemblies.

     Several  microwave navigation systems with  sufficient accuracy and
adequate range are available for $40,000-$90,000.   These systems  comprise
two  shore  stations and an on-board transmitter.  With an additional  shore
station, the hyperbolic mode  can provide multiple user capability.  Limitations
include geometry of shore stations;  vessel position in the coverage area
(i.e.,  crossing angle limitations); and possible  interferences from line-of-site
obstructions,  sea-surface  reflective nulls, and land-sea boundaries.

     Positioning  from Loran-C ranges offers acceptable repeatable  accuracy
for relocating at  station  locations for most  sampling objectives.   Receivers
cost  $1,000-$4,000 and do not  require additional personnel.  Limitations
include interference in  some areas of Puget Sound and the need to  initially
locate  stations with another method.

     Transit  satellite-based  methods currently  do  not offer sufficient
accuracy except with multiple passes, which  are  impracticable when  a sampling
station is only briefly  occupied.  In the future,  required accuracies will
be achievable using Navstar GPS satellite-based techniques ($10,000-$40,000
for  first units; $1,000  for subsequent  production models).  Independent
geosynchronous satellite networks,  such as GEOSTAR, may become  available
at  a  system  interrogator cost  of  $450 plus  a  monthly fee.  This method
is in the early planning  stages and recently  received FCC  approval.   Satellite
methods do not require  additional personnel.

Range  and Angle

     A theodolite and EDMI  could be paired with  a communications link  for
approximately $10,000-$12,000.  Total stations developed for this  purpose
range in cost from $9,000 for  a  manual  station  to  $15,000-$25,000  for a


                                    D-5

-------
                                                       Station Positioning
                                            Appendix D - Methods Evaluation
                                                               August 1986


fully automated  station.  Optical and infrared  range limitations exist, and
the optical  components can be operated only during daylight hours  of good
visibility.   The  range-azimuth positioning methods  examined  (see Appendix C)
provide  sufficient  positional accuracy with a single  station, and cost
between  $65,000 for  a manual tracking and $70,000-$100,000 for fully automated
tracking.

SCREENING  CRITERIA

     Candidate systems were  evaluated  for  accuracy, flexibility  (i.e.,
range of conditions  under which the system can operate,  including  use for
other purposes),  portability,  reliability, servicing requirements,  availability,
cost, and convenience.  Results of the evaluation are  presented in Table D-3.
Methods are presented  across the top of this  table in order of increasing
range capability.  Methods that  could be eliminated from  further consideration
for some  sampling  objectives are marked by an asterisk.   Limitations that
could preclude further consideration included inability to operate at  night
or in poor  visibility (e.g.,  optical methods) or availability of comparable
methods  at lower  cost with  fewer  logistical problems  (e.g., medium-range
systems).   The investigator has to determine the project-specific considerations
that could be restrictive enough  to preclude use of some positioning methods
and to  determine if the accuracies required for the study  can  be achieved.
For purposes of  this evaluation, medium-range sytems  were dropped from
consideration and optical methods were considered only  for nearshore sampling.
Remaining positioning  methods were  reevaluated for range  capability, accuracy,
availability, capital and operating costs, and merits of use.  This information
is summarized below.

Range Considerations

     Most of the systems  noted  for restrictive limitations have inadequate
range capabilities  for  visibility  and/or  distance  reasons.   At  sampling
sites farther from  land  in  Puget  Sound, VRR will have to  be  used at  a  larger
scale and  a  lower  accuracy.  Sextants and  (to a  lesser  extent)  systems
that  use  theodolites may  not permit adequate resolution of targets.   Lower
resolution results  in reduced  accuracy.  The rest of the remaining  systems
listed  in Table  D-3 have adequate ranges to cover most areas of Puget Sound
(Figure D-l).

Accuracy

     Various methods can  meet  repeatable accuracies of  +20  m (+66 ft).
VRR can consistently position  within +20  m  (+66 ft) in  areas close  to  shore
with  supplement from  visual  fixes and  farther from land in areas with good
Loran-C reception.   Loran-C  alone  can  reposition within +20 m  (+66 ft)
in  the  Elliott  Bay area that has  recently been mapped with a  Loran  signal
grid.  Also, under  good  conditions,  sextants  can easily reposition  within
this  distance.  Near shore,  sextants may approach the accuracy  of microwave
systems under perfect conditions. Microwave systems, range-azimuth  systems,

                                    D-6

-------
                            TABLE  D-3.   EVALUATION OF  NAVIGATION  METHODS FOR STATION POSITIONING
Sextant
Angle
Resection Theodolite
Accuracy
Absolute
Repea table
Flexibil ity
Portabil ity
Reliabil ity
Servicing
Cal ibration
Maintenance
Availabil ity
Equ i pmen t
Service
Rental
Cost
Convenience

M-L
H-M
M-L*
H
H
L
H

H
M-L
H
H
L

H-M
H
L*
M
H-M
L
M

H
M
H
H
L
Variable
Range
Radar
(VRR)

L
M
H
L
H-M
H
H

H
H
L
H-M
H
VRR,
Visuals,
Loran-C

M-L
H-M
H
H-L
H-M
H-M
H

H
H
H-L
H-M
H
Total
Stations

H-M
H
M-L*
M
H-M
M-L
M

n
M-L
M-L
M-L
L
Microwave
Systems

H
H
H-M
H
H-M
L
H

H
M
H
L
L
Range- Medium-
Azimuth Range
Systems Systems Loran-C

H-N
H
M-L*
H
H-M
L
M

M-L
M-L
L
L*
L

H-M
M
M-L
L
H-M
H-M
H-M

L*
M-L
L
L*
L*

M
M
H-L
H
H
H-M
H

H
H-M
H
H
H
GEUbTAK GPi

H h-11
H H-M
H H
H-M H-L
? M
H H
H H

? M
* M
? '
H-L M-L
H H-L
H = High ranking (adequate, above average, inexpensive, infrequent).
M = Medium ranking (marginal, average,  intermediate).
L = Low ranking (not adequate, below average, expensive, frequent).
• - Significant  enough  limitation to preclude use as a positioning method
under some conditions.
' = Not enough information available to evaluate.

-------
oo
                        3
                        o
                        <
                        2
                        Q-
                        LU
                        a:
                            100 —
                            120
                                                                                  MICROWAVE SYSTEMS

                                                                                  AND SOME RANGE-ANGLE

                                                                                  SYSTEMS
                                                                       PUGET SOUND

                                                                       MAXIMUM RANGE
                                                10               20



                                             DISTANCE FROM SHORE (TARGETS) (km)
                    Figure D-l.   Approximate accuracy versus distance of some  positioning  methods.

-------
                                                       Station Positioning
                                           Appendix D - Methods Evaluation
                                                              August 1986


total  stations,  and theodolites can reposition a vessel within +2 m (+6.6 ft).
The accuracies of  the latter three methods will deteriorate with the visi-
bility.   The  GPS satellite method accuracies, which vary  with satellite
code access,  are  between  those of microwave  systems and the repeatable
accuracy  of  Loran-C and  VRR.   GEOSTAR will be almost as accurate as the
microwave  systems.

     The only systems  that are  capable of obtaining  absolute accuracies
greater than  +20 m (+66 ft) in most locations  are  the electronic  ranging
systems (microwave-based or  satellite).  Near shore, optical methods and
VRR can usually  reach these absolute accuracies  but are dependent on the
available  targets  and calm, clear conditions.

Availability

     The satellite systems will  not  be usable for an undetermined period
of time. GPS  does  not yet  have enough satellites in orbit to give consistent
fixes  without long time  delays.  GEOSTAR will not be operational for  a
few years. The  other candidate systems are  available.

Capital and Operating Costs

     GPS capability is expected to cost from $10,000 to $40,000.  Cost
of later models is expected  to drop to $1,000.  Proposed GEOSTAR interrogators
are expected to  cost  $450 plus a monthly use fee.  VRR models range from
$4,000  to $10,000.   Loran-C units cost $1,000-$4,000.  Little or no operating
cost  is associated with  these methods  and the vessel captain can operate
them.  Microwave  systems  typically cost $40,000-$95,000,  not  including
high operating expenses.   Range-azimuth systems range  from $64,500 to $100,000
and offer greater flexibility for other work  and reduced logistical problems.
Remaining  systems (total stations for $9,000 manual, $15,000-$25,000 automatic;
EDMI and theodolites for  $10,000-$12,000; and  pairs of  sextants or theodolites
for $4,000-$8,000) are all  significantly  lower in  initial cost.   However,
these are  typically much more labor-intensive and logistically complicated,
and can have  high operating costs.

     Short-term rental  or  long-term leasing  of equipment should be considered
as an alternative to purchase of  a positioning  system.  A short-term rental
limits cost  to only the  actual surveying periods and avoids maintenance
responsibilities.   These arrangements  are also attractive during periods
of  rapid  navigation system development (e.g., satellite systems) if such
developments  indicate that adequate systems  will  be  available in   the near
future at  lower capital costs.

     The  monthly rental  cost  of equipment is often set at 10  percent  of
the suggested retail cost. However,  short-term rates  for many items are
categorized  by the number of rental days (e.g.,  1-10,  11-90 and  91 plus).
Rent/purchase options  are available with a set amount of the rental  fee
(e.g., 50-80 percent) applied  to  the purchase price  when the  option  is

                                   D-9

-------
                                                      Station Positioning
                                           Appendix D - Methods Evaluation
                                                              August  1986


exercised.   Typical  rental  costs  (subject to change) for items of interest
are shown  in  Table D-4.  Evaluation of the cost-effectiveness of applicable
systems should  include rent/lease options as well as procurement.   If the
user has little experience in a  particular system, it may  also be appro-
priate to  include the cost of one or more equipment  technicians or operators
(up to $300  per  day plus  expenses)  to provide reliable  navigation  data
for each survey.

Merits of  Use

     When  repeatable  positioning is the primary concern, both Loran-C and
VRR methods  have adequate accuracies for most sampling objectives.  A combi-
nation of VRR,  visual fixes,  Loran-C,  and recording  station depth  will
improve repeatable accuracy.   Both  Loran-C and VRR are  routinely carried
on research  vessels, limiting logistical requirements.  The combined method
also offers  redundancy that minimizes the effect on  positioning capability
if one of  the systems malfunctions.

     Sextant  fixes  are affected by distance, visibility, and rough water.
The variability in positioning accuracy across the  range of conditions
that can be  expected make this method questionable for consistent operations
across all  stations.  In good conditions, sextant fixes can improve accuracies
over  those  obtainable for  VRR or Loran-C.  However,  if the investigator
is trying  to locate specific stations, targets and  angles  have to be  pre-
determined.   Sextants are  also valuable as a backup  positioning method
when the primary system malfunctions.

     The  shore-based  optical methods are  not  as  affected by rough water.
Logistical requirements  include location of the  shore stations and one
or two shore station operators.   Range-azimuth and total stations offer
the advantages of a single  shore  station and  a built-in communications
link, which will  limit  logistical  requirements.  Two  stations have to be
staffed and a three-way communications link has to be provided for theodolite
positioning.  These systems  can  achieve accuracies of approximately +2 m
(+6.6 ft), but accuracies can deteriorate  with visibility and distance
to shore for the optical components of  these  systems.

     Microwave  systems offer  high accuracies [+2 m (+6.6 ft)], but at a
high cost.  Microwave  systems are not limited  by visibility restrictions
or ranges  in Puget Sound.  The additional  logistical requirements of microwave
positioning  add  complexity to the study  design  and implementation.   When
available (approximately 1988),  consistent satellite fixes are expected
to offer the increased accuracies of microwave  sytems  with the  simplicity
of Loran-C.

     The  radius of  the probable  sampling area could  be  reduced by  <50 m
(165 ft) by using a  +2-m  absolute accuracy  system rather  than  a +20-m absolute
accuracy  system.   In deeper waters (e.g.,  increased  sampling  area but more
homogeneous   benthos),  the added accuracy  would only be advantageous if

                                   D-10

-------
         TABLE 0-4.   TYPICAL EQUIPMENT RENTAL COSTS (1985 DOLLARS)
Category
Theodolites
EOMIs
Total Stations


1"
6-10"
20"
1.0-2.0 km
2.0-6.0 km
>6.0
Automatic
Manual
1-10 Days
25-35
22-25
15-18
25-30
25-65
40-50
75-110
40-50
Rental Period
11-90 Days 91+ Days Monthly
18-21
13-15
10-13
15-19
18-39
25-30
45-70
25-45
12-15
9-12
7-10
10-13
15-20
16-22
30-60
16-20
Hydrographic Systems

  Trisponder 202                    175                             2,750

  Cubic Auto Tape
    (3 range)                       200         105          90

  Motorola Mini
    Ranger II
    (including 2 trans-
    ponders)                        200         105          90     3,750
    (additional trans-
    ponders)                         36          18          12
                                     D-ll

-------
                                                      Station Positioning
                                           Appendix  0 - Methods Evaluation
                                                              August 1986


station grid  size or absolute positions were critical.  In shallower, near
shore waters,  the differences in achievable accuracy decrease as resolution
of  positioning targets  improves.  However, spatial  heterogeneity also
increases.   The investigator will have to determine  if  the increase  in
probable sampling area [now less  than 30 m (98 ft) in radius] that results
from using  less accurate systems will compromise  quality  of the collected
data.  If stations were initially located using a highly accurate positioning
method (e.g.,  microwave system) and position coordinates were simultaneously
defined in  coordinates of  a lower accuracy method (e.g.,  Loran-C, VRR,
visual fixes), the station  could be reoccupied using  the less  accurate
method, without significant loss in absolute accuracy.
                                   D-12

-------
SEDIMENT CONVENTIONAL*

-------
FINAL REPORT                     --_,..
TC-3991-04                        Puget Sound Estuary Program
RECOMMENDED PROTOCOLS FOR MEASURING
CONVENTIONAL SEDIMENT VARIABLES
IN PUGET SOUND
Prepared by:

TETRA TECH, INC.
Prepared for:
U.S. ENVIRONMENTAL PROTECTION AGENCY
Region 10 - Office of Puget Sound
Seattle, WA
March, 1986
TETRA TECH, INC.
11820 Northup Way
Bellevue, WA 98005

-------
                                 CONTENTS


                                                                        Page

LIST OF TABLES                                                            v

INTRODUCTION                                                              1

COLLECTION OF SURFICIAL SEDIMENTS FOR PHYSICAL AND CHEMICAL
VARIABLES                                                                 4

     INTRODUCTION                                                         4
     DESIGN OF SAMPLER                                                    4
     PENETRATION DEPTH                                                    5
     OPERATION OF SAMPLER                                                 5
     SAMPLE ACCEPTABILITY CRITERIA                                        6
     SAMPLE COLLECTION                                                    6

PARTICLE SIZE                                                             9

     USE AND LIMITATIONS                                                  9
     FIELD PROCEDURES                                                     9

          Collection                                                      9
          Processing                                                     10

     LABORATORY PROCEDURES                                               10

          Analytical Procedures                                          10
          QA/QC Procedures                                               15

     DATA REPORTING REQUIREMENTS                                         16

TOTAL SOLIDS                                                             I7

     USE AND LIMITATIONS                                                 17
     FIELD PROCEDURES                                                    17

          Collection                                                     17
          Processing                                                     17

     LABORATORY PROCEDURES                                               17

          Analytical Procedures                                          17
          QA/QC Procedures                                               I8

     DATA REPORTING REQUIREMENTS                                         19
                                       ii

-------
TOTAL VOLATILE SOLIDS (TVS)                                              20
     USE AND LIMITATIONS                                                 20
     FIELD PROCEDURES                                                    20
          Collection                                                     20
          Processing                                                     20
     LABORATORY PROCEDURES                                               20
          Analytical Procedures                                          20
          QA/QC Procedures                                               21
     DATA REPORTING REQUIREMENTS                                         22
TOTAL ORGANIC CARBON (TOC)                                               23
     USE' AND LIMITATIONS                                                 23
     FIELD PROCEDURES                                                    23
          Collection                                                     23
          Processing                                                     23
     LABORATORY PROCEDURES                                               23
          Analytical Procedures                                          23
          QA/QC Procedures                                               25
     DATA REPORTING REQUIREMENTS                                         25
OIL AND GREASE  (FREON EXTRACTABLE)                                       27
     USE AND LIMITATIONS                                                 27
     FIELD PROCEDURES                                                    27
          Collection                                                     27
          Processing                                                     27
     LABORATORY PROCEDURES                                              28
          Analytical Procedures                                          28
          QA/QC Procedures                                              30
     DATA REPORTING REQUIREMENTS                                         31
TOTAL  SULFIDES                                                           32
     USE AND  LIMITATIONS                                                 32
     FIELD  PROCEDURES                                                    32
          Collection                                                     32
          Processing                                                     32

-------
     LABORATORY PROCEDURES                                               32

          Analytical Procedures                                          32
          QA/QC Procedures                                               35

     DATA REPORTING REQUIREMENTS                                         36

TOTAL NITROGEN                                                           37

BIOCHEMICAL OXYGEN DEMAND (BOD)                                          38

     USE AND LIMITATIONS                                                 38
     FIELD PROCEDURES                                                    38

          Collection                                                     38
          Processing                                                     38

     LABORATORY PROCEDURES                                               38

          Analytical Procedures                                          38
          QA/QC Procedures                                               41

     DATA REPORTING REQUIREMENTS                                         41

CHEMICAL OXYGEN DEMAND (COD)                                             42

     USE AND LIMITATIONS                                                 42
     FIELD PROCEDURES                                                    42

          Collection                                                     42
          Processing                                                     42

     LABORATORY PROCEDURES                                               43

          Analytical Procedures                                          43
          QA/QC Procedures                                               45

     DATA REPORTING REQUIREMENTS                                         45

REFERENCES                                                               46
                                     iv

-------
                              LIST OF TABLES


Number                                                                  Page

   1    Contributors to the sediment  conventional  protocols                2

   2    Recommended sample sizes, containers,  preservation,
        techniques, and holding  times for sediment conventional
        variables                                                         3

   3    Withdrawal times for pipet analysis as a function
        of particle size and water temperature                           14
                             ACKNOWLEDGEMENTS


     This chapter was  prepared  by  Tetra Tech, Inc., under  the  direction  of
Dr. Scott Becker,  for the  U.S.  Environmental  Protection Agency  in  partial
fulfillment of  Contract  No.  68-03-1977.   Dr.  Thomas Ginn of Tetra Tech was
the Program Manager.   Mr. John  Underwood  and  Dr.  John  Armstrong of U.S. EPA
were the  Project  Officers.   Ms. Julia  Wilcox  of Tetra Tech helped  resolve
many  of  the technical  issues  that  arose  as the  protocols were  being
prepared.

-------
                                            Conventional Sediment Variables
                                                               Introduction
                                                                 March 1986
                               INTRODUCTION
     Recommended methods for measuring  the  following  conventional  sediment
variables in Puget Sound are  presented in this chapter:

          Particle size
          Total  solids
          Total  volatile solids
          Total  organic carbon
          Oil and grease
          Total  sulfides
          Total  nitrogen
          Biochemical oxygen  demand
          Chemical oxygen demand.

Each method  is  based  on the results  of a  workshop and written  reviews  by
representatives  from most organizations  that fund  or  conduct  environmental
research in  Puget  Sound  (Table  1).   The  purpose of  developing  these
recommended  protocols  is to  encourage all  Puget  Sound  investigators
conducting   monitoring programs, baseline  surveys,   and  intensive  invest-
igations to  use standardized methods whenever  possible.   If  this goal  is
achieved, most  data collected in  Puget  Sound  should  be directly  comparable
and thereby  capable of  being integrated  into a sound-wide  database.   Such a
database is  necessary  for  developing  and maintaining  a comprehensive water
quality management program for Puget Sound.

     Each  recommended   protocol   describes  the use  and limitations  of  the
respective variable; the  field  collection  and processing  methods; and  the
laboratory   analytical,  QA/QC,  and  data  reporting procedures.   Each
recommended   analytical  procedure was modified  from Plumb  (1981).   The
general  collection and holding recommendations  for each variable  are
presented in Table 2.

     Although   the  following  protocols are  recommended  for most  studies
conducted in Puget Sound, departures from these methods may be necessary to
meet the special requirements  of individual  projects.  If such  departures
are made, however, the  funding agency or investigator should  be  aware that
the resulting data may  not be comparable with most other data of that kind.
In some instances, data collected using  different methods may be compared if
the methods  are intercalibrated  adequately.

-------
     TABLE 1.   CONTRIBUTORS TO THE SEDIMENT CONVENTIONAL PROTOCOLS
             Name
        John Armstronga
        Bob Barricka
        Scott Becker*»b
        Gordon Bilyard*
        Chuck Boatman*
        Eric Crecelius3
        Joe Cummins
        Bob Dexter3
        George Ditsworth
        John Downing9
        Bruce Duncan
        Mark Fugiel*
        Arnold Gahler
        Carolyn Gangmark
        Roy Jones
        Dave Kendall
        Shawn Moore*
        Gary Mauseth*
        Mike Nelson*
        Ahmad Nevissi*
        Wally Triol*
        Frank Urabeck
        Steve Vincent*
        Fred Weinmann*
        Julia Wilcoxa
        Carolyn Wilson
        Jack Word*
Organization
U.S. EPA
Tetra Tech, Inc.
Tetra Tech, Inc.
Tetra Tech, Inc.
URS Engineers
Battelle Northwest
U.S. EPA
EVS Consultants
U.S. EPA
Nortec
U.S. EPA
Am Test, Inc.
U.S. EPA
U.S. EPA
U.S. EPA
U.S. COE
Am Test, Inc.
Nortec
Laucks Testing Labs
Univ. of Washington
Parametrix, Inc.
U.S. COE
Weyerhauser Company
U.S. COE
Tetra Tech, Inc.
U.S. EPA
Evans-Hamilton,  Inc.
Attended the workshop held on June 28, 1985.
Workshop moderator.

-------
 TABLE 2.  RECOMMENDED SAMPLE SIZES,  CONTAINERS,  PRESERVATION TECHNIQUES,
           AND  HOLDING TIMES  FOR SEDIMENT CONVENTIONAL VARIABLES
Minimum
Sample
Variable Size (g)*
Particle size
Total solids
Total volatile
solids
Total organic
carbon
Oil and grease
Total sul fides
Total nitrogen
Biochemical
oxygen demand
Chemical oxgyen
demand
100-15QC
50
50
25
100
50
25
50
50
Contained
P,G
P.G
P,G
P,G
G only
P,G
P,G
P,G
P.G
Preservation
Cool, 4° C
Freeze
Freeze
Freeze
Cool, 4° C, HC1;
Freeze
Cool, 40 C,
IN zinc acetate
Freeze
Cool, 40 C
Cool, 40 C
Maximum
Holding
Time
6 mod
6 mod
6 mod
6 mod
28 daysd
6 mod
7 daysd
6 mod
7 dayse
7 days6
a Recommended field sample sizes  for  one  laboratory analysis.  If additional
laboratory analyses are required  (e.g., replicates), the  field sample  size
should be adjusted accordingly.
b p = polyethylene, G = glass.
c Sandier sediments require larger  sample sizes  than do muddier sediments.
^ This  is a suggested holding  time.   No U.S.  EPA criteria exist for the
preservation of this variable.
e This holding time is recommended  by Plumb  (1981).

-------
                                             Conventional Sediment Variables
                                              Surficial Sediment Collection
                                                                 March 1986
                     COLLECTION OF  SURFICIAL  SEDIMENTS
                    FOR PHYSICAL AND CHEMICAL VARIABLES
INTRODUCTION
     This section describes the protocols required to collect an acceptable
subtidal surficial  sediment  sample for subsequent  measurement  of physical
and chemical  variables.   This  subject has generally  been neglected  in the
past and sampling crews have been  given relatively wide  latitude  in deciding
how  to  collect  samples.   However,   because  sample  collection procedures
influence the results of all subsequent laboratory and data analyses,  it  is
critical that  samples be  collected using  acceptable and standardized
techniques.

DESIGN OF SAMPLER

     In Puget Sound, the most common  sampling  device  for  subtidal  surficial
sediments is  the modified  van  Veen  bottom  grab.  However,  a  variety  of
coring devices is also  used.  The  primary criterion for an  adequate  sampler
is that  it  consistently collect undisturbed samples  to the required  depth
below the sediment surface without contaminating  the samples.  An additional
criterion is  that the  sampler  can be  handled  properly  on board the survey
vessel.   An  otherwise acceptable  sampler  may  yield  inadequate sediment
samples if it is too large, heavy, or  awkward to  be  handled  properly.

     Collection of undisturbed sediment requires  that  the  sampler:

     •    Create a minimal bow wake when descending
     •    Form a leakproof seal  when the sediment sample is  taken
     •    Prevent  winnowing and excessive  sample  disturbance  when
          ascending
     •    Allow easy access to the sample surface.

Most modified van Veen grabs  have open  upper faces  that  are  fitted  with
rubber flaps.  Upon  descent,  the  flaps are  forced open to  minimize  the bow
wake, whereas  upon  ascent, the flaps are forced closed  to prevent sample
winnowing.   Some  box  corers  have  solid  flaps  that  are  clipped open  upon
descent and snap shut after the corer is  triggered.   Although most samplers
seal  adequately  when  purchased,  the  wear  and  tear  of repeated field use
eventually  reduces  this  sealing  ability.   A  sampler should  therefore  be
monitored constantly  for  sample  leakage.   If  unacceptable leakage  occurs,
the sampler should be repaired or  replaced.  If  a sampler is to  be borrowed
or  leased  for a project,  its  sealing ability  should be  confirmed prior  to
sampling.  Also, it is prudent to  have a  backup  sampler on  board the survey
vessel in case the primary sampler begins leaking during a cruise.

-------
                                             Conventional Sediment Variables
                                              Surficial Sediment Collection
                                                                 March 1986
     The required penetration depth  below the  sediment  surface  is a function
of  the  desired  sample  depth  (see  Penetration  Depth).   Generally,  it  is
better  to  penetrate below  the desired  sample  depth  to  minimize  sample
disturbance  when the  sampling device  closes.   Penetration depth  of  most
sampling devices  varies with  sediment  character,  and generally is greatest
in fine sediments and least in coarse sediments.   Sampling  devices generally
rely upon either  gravity  or a  piston  mechanism  to  penetrate the sediment.
In  both  cases,  penetration depth can  be modified by adding or subtracting
weight from the samplers.  Thus, it  is  optimal to use  a sampler that has a
means of weight adjustment.  If a  sampler  cannot consistently achieve the
desired penetration  depth, an alternate device should be used.

     Once the sampler is secured on  board the  survey  vessel,  it  is essential
that the surface  of the  sample be  made  accessible  without  disturbing the
sample.  Generally,  samplers have hinged  flaps on their  upper  face for  this
purpose.  The opem'ng(s)  in  the upper face of the  sampler should be  large
enough to allow easy subsampling of  the sediment  surface.  If  an  opening  is
too  small,  the sample  may  be  disturbed as  the scientific  crew  member
struggles to take a  subsample.

PENETRATION DEPTH

     For characterizing  surficial   sediments  in  Puget  Sound,   it  is
recommended that  the upper  2  cm of  the sediment  column  be evaluated.   When
collecting  the  upper 2  cm of  sediment,  it  is  recommended  that  a  minimum
penetration depth of 4-5 cm be achieved for each  acceptable sample.

     Although the 2-cm specification is arbitrary, it will  ensure  that:
     •    Relatively recent sediments are sampled
     •    Adequate  volumes  of  sediments  can  be  obtained  readily for
          laboratory analyses
     •    Data from different studies can be compared validly.

Sampling depths other than  2 cm may  be appropriate for specific purposes.
For example, the  upper 1 cm of sediment may be required to  determine  the age
of the most recently deposited  sediments.   By contrast,  a  sample  depth  much
greater than  2  cm may  be required  to evaluate the  vertical  profile  of
sediment characteristics  or  to determine  depth-averaged  characteristics
prior to dredging.  If a sampling depth other  than 2 cm is  used, comparisons
with data from 2-cm deep samples may be questionable.

OPERATION OF SAMPLER

     The sampling device  should be  attached to the  hydrowire  using  a ball-
bearing swivel.  The swivel will minimize the  twisting  forces on the  sampler
during  deployment and  ensure that  proper contact is made with the  bottom.

-------
                                             Conventional  Sediment Variables
                                               Surficial Sediment Collection
                                                                 March  1986
For  safety,  the hydrowire,  swivel,  and  all  shackles  should  have  a  load
capacity at least three times greater than the weight of a  full  sampler.

     The sampler should be lowered through the water column at  a  controlled
speed of approximately 1 ft/sec.  Under no circumstances should the  sampler
be allowed  to  "free fall" to  the  bottom, as this  may  result   in  premature
triggering, an excessive bow wake,  or improper orientation  upon  contact with
the  bottom.   The sampler  should contact  the bottom gently  and  only  its
weight or piston mechanism should be used  to  force it into  the sediment.

     After the sediment sample is taken, the  sampler should be raised slowly
off  the  bottom  and  then retrieved at a controlled  speed of approximately  1
ft/sec.  Before the sampler  breaks  the  water  surface, the  survey vessel
should head  into the waves  (if  present)  to  minimize vessel rolling.  This
maneuver will minimize  swinging  of  the sampler  after  it  breaks  the water
surface.  If excessive swinging  occurs or  if  the  sampler strikes  the vessel
during  retrieval,  extra attention  should be paid  to  evaluating  sample
disturbance when judging sample acceptability.

     The sampler should be secured immediately after  it  is  brought on board
the  survey  vessel.   If  the  sampler tips  or  slides  around  before being
secured, extra attention should be paid  to evaluating sample disturbance.

SAMPLE ACCEPTABILITY CRITERIA

     After  the  sampler is  secured  on deck,  the  sediment  sample  should  be
inspected  carefully before  being  accepted.   The  following acceptability
criteria should be satisfied:

     •    The  sampler  is  not  over-filled  with  sample  so  that  the
          sediment surface is pressed against the  top of the sampler
     •    Overlying water is present (indicates  minimal  leakage)
     •    The  overlying  water  is   not excessively turbid  (indicates
          minimal sample disturbance)
     •    The sediment  surface  is  relatively flat  (indicates  minimal
          disturbance or winnowing)
     t    The desired penetration depth  is achieved (i.e.,  4-5 cm  for a
          2-cm deep surficial sample).

If a sample does not meet all criteria,  it should  be rejected.

SAMPLE COLLECTION

     After  a sample is  judged  acceptable,  the following observations should
be noted on the field log sheet:

     •    Station location
     •    Depth

-------
                                             Conventional Sediment Variables
                                               Surficial Sediment Collection
                                                                 March 1986
     •    Gross characteristics  of the surficial  sediment
               Texture
               Color
               Biological structures (e.g.,  shells,  tubes,  macrophytes)
               Presence of debris (e.g.,  wood  chips, wood fibers, human
               artifacts)
               Presence of oily  sheen
               Odor (e.g., hydrogen sulfide, oil, creosote)
     •    Gross characteristics  of the vertical  profile
               Changes in sediment characteristics
               Presence  and  depth  of  redox potential discontinuity
               (rpd) layer
     •    Penetration depth
     •    Comments related to sample quality
               Leakage
               Winnowing
               Disturbance.

     Before subsamples of the surficial  sediments are  taken,  the  overlying
water must be  removed.   The preferred method of  removing  this water is  by
slowly  siphoning  it  off near  one  side  of  the  sampler.   Methods such  as
decanting the water or slightly  cracking the  grab to  let the water run  out
are not recommended,  as  they  may result  in  unacceptable disturbance  or  loss
of fine-grained surficial sediment and organic matter.

     Once the  overlying  water has been removed,  the surficial  sediment  can
be subsampled.   It is recommended  that  subsamples be  taken  using  a  flat
scoop shaped like a coal  shovel.   The shoulders  of  the  scoop should  be  2 cm
high.   This  device will  allow   a  relatively  large subsample to  be  taken
accurately to  a  depth of 2 cm.   Coring  devices  are not recommended  because
generally they  collect small amounts of surficial sediment and  therefore
require repeated  extractions  to obtain a sufficient volume  of  material  for
analysis  of  conventional  sediment  variables.  A  curved  scoop  is   not
recommended because  it does  not  sample  a  uniform depth.   Because accurate
and  consistent subsampling  requires  practice,   it is  advisable that  an
experienced person perform this  task.

     When subsampling surficial  sediments,   unrepresentative material  should
be removed  in  the field  under  the supervision  of  the  chief  scientist  and
noted on the field log sheet.  The criteria used to determine representa-
tiveness should be determined prior to sampling.

     Finally,  if  samples  are to be  analyzed  for trace metals or priority
pollutant organic compounds,  sample contamination during collection must be
avoided.  All  sampling  equipment  (i.e., siphon   hoses,  scoops,  containers)
should  be  made of  noncontaminating material  and  should  be  cleaned  appro-
priately before  use.   Samples should not be  touched with  ungloved fingers.
In addition, potential airborne  contamination  (e.g., stack  gases,  cigarette
smoke)  should  be avoided.   Detailed  guidance  for  preventing  sample

-------
                                             Conventional Sediment Variables
                                               Surficial Sediment Collection
                                                                  March 1986
contamination is given in the protocols for metals  and  organic  compounds in
other chapters of this notebook.
                                       8

-------
                                            Conventional Sediment Variables
                                                              Particle Size
                                                                 March 1986
                               PARTICLE  SIZE
USE AND LIMITATIONS

     Particle size  is  used to characterize the physical characteristics of
sediments.  Because particle  size influences both  chemical  and biological
variables, it can be used  to  normalize  chemical concentrations according to
sediment characteristics and to account  for some of  the  variability  found in
biological assemblages.   Particle size  is also  an important variable for
marine  engineering  purposes.    In  addition to  Plumb (1981),  a  variety of
other  references  discuss  the uses and  measurement  of particle size  (e.g.,
Krumbein and Petti John 1938; Folk 1968;  Buchanan 1984).

     Particle size  can be characterized  in  a  wide  range  of detail.  The
grossest divisions  that generally  are considered  useful for  characterizing
particle  size  distributions are percentages  of  gravel,  sand,  silt, and
clay.   However, each  of these size  fractions can be subdivided further so
that  additional  characteristics  of  the  size distribution (e.g.,  mean
diameter, skewness, kurtosis) can be determined.

     Particle  size determinations  can  either include  or exclude  organic
material.   If organic  material  is  removed  prior  to  analysis,  the  "true"
(i.e.,  primarily  inorganic)  particle size distribution is determined.   If
organic material  is included  in  the  analysis, the "apparent11  (i.e., organic
plus  inorganic) particle  size distribution is determined.   Because  true and
apparent  distributions may  differ,  detailed comparisons  between  samples
analyzed  by  these different  methods  are  questionable.   It is  therefore
desirable  that  all samples within  each  study (at a  minimum)  and  among
different  studies  (if  possible) be  analyzed  using  only  one of these two
methods.

FIELD  PROCEDURES

Collection

      Samples  can  be  collected in glass  or plastic  containers.  A minimum
sample size of  100-150  g  is recommended.   If unrepresentative material is to
be removed from  the  sample,  it  should be  removed in the  field  under the
supervision of  the  chief  scientist and noted on the field log sheet.

-------
                                             Conventional  Sediment Variables
                                                              Particle Size
                                                                 March 1986
Processing
     Samples should be stored at 4° C, and can  be held  for  up  to  6 mo  before
analysis.  Samples must not be frozen or dried prior to analysis, as  either
process may change the particle size distribution.

LABORATORY PROCEDURES

Analytical Procedures

•    Equipment
          Sieve shaker
               Ro-Tap or equivalent
          Drying oven
          Constant temperature bath
          Analytical balance
               0.1 mg accuracy
          Desiccator
          Clock
               With second hand
          Standard sieves
               Appropriate mesh sizes
          Sieve pan and top
          Sieve brush
          Funnel
          1-L graduated cylinders
          50-mL beakers
          20-mL pipets
          Water pique or squirt bottle
          Glossy paper
          Dispersant
               1  percent  sodium hexametaphosphate = 1 percent  commercially
               available Calgon
          Distilled water.

0    Sample preparation
     -Allow samples to warm to room temperature.
          Homogenize each  sample mechanically,  incorporating any overlying
          water.
          Remove  a  representative  aliquot (approximately 25 g)  and analyze
          for  total  solids  content.   This  information  can  be used to
          estimate  the dry  weight  of  the  aliquot used  for  particle  size
          analysis.  The  efficiency  of  the entire  analysis  can then be
          evaluated by  adding the dry weights of all  sample fractions  and
          comparing this  sum with  the estimated  dry weight of the original
          aliquot.
          Remove  a  second  representative  aliquot for  wet  sieving.   The
          aliquot can range from 20 g for muddy sediments  to 100  g for sandy
          sediments.   The  critical  factor for sample  size determination  is

                                      10

-------
                                       Conventional Sediment Variables
                                                         Particle Size
                                                            March 1986
     the weight  of fine-grained  material  that will  be used  for  the
     pipet  analysis.   Ideally  the  total  dry weight  of fine-grained
     material in the 1-L graduated cylinder should equal approximately
     15 g.   However,  total  weights between 5  and  25 g are considered
     acceptable.  Total weights outside this  range  are not considered
     acceptable and it is recommended that aliquot size be modified to
     bring the  amount of  fine-grained  material  into  the acceptable
     range.
     Weigh the wet sample to the nearest 0.01  g.

Organics  oxidation  -  this  step  removes  organic material  from  the
sample.It is optional and depends upon the objectives  of each study.
     Place the sediment sample in  a large  beaker (>_2 L).
     Add 20 ml of 10 percent hydrogen  peroxide solution  and mix.
     Let the sample stand until  frothing stops.
     Once frothing stops, add an additional 10 ml of hydrogen peroxide
     solution.
     Continue adding 10-mL portions of hydrogen peroxide solution until
     no frothing occurs on addition.
     Boil the sample to remove any excess  hydrogen  peroxide.
     Be  careful  that material  is  not  lost from the beaker during
     frothing and boiling.

Wet-sieving  -  this  step  separates  the sample  into size fractions
greater  than  62.5  urn  (i.e.,  sand and gravel)  and less  than  62.5 urn
(i.e., silt and clay)
     Place the 62.5-um (4 phi)  sieve  in a  funnel, with  a 1-L graduated
     cylinder underneath.   Moisten the sieve  using  a  light  spray of
     distilled water.
     Place the sample in a beaker, add 20-30 mL of  distilled water, and
     stir to suspend fine-grained  material.
     Pour the sample into the sieve and thoroughly  rinse the beaker and
     stirrer with distilled water.
     Wash the sediment on the sieve with  distilled water using a water
     pique or squirt bottle having low water  pressure.  Aggregates can
     be gently broken using a rubber policeman.
     Continue wet  sieving  until  only clear  water  passes  through  the
     sieve.   Try  to  ensure  that  the  rinsate  does  not  exceed
     approximately  950  mL.   This  can  generally be  accomplished by
     sieving a sample quantity  that  is not too large and by efficient
     use  of  the  rinse  water.  Both  of  these techniques  may require
     experimentation before routine wet sieving is  started.

Gravel-sand  fraction  -  this fraction  is   subdivided  further  by
mechanically dry sieving it through a graded series of  screens.
     Wash  the  coarse  fraction into  a preweighed 50-mL  beaker using
     distilled water.   Rinse the sieve thoroughly.
                                 11

-------
                                       Conventional Sediment Variables
                                                         Particle Size
                                                            March 1986
     Dry  the  coarse  fraction to  constant  weight at  90  +_ 2°  C.   The
     drying temperature  is  less  than  100° C  to prevent  boiling and
     potential  loss of  sample.
     Cool  the sample  to room temperature  in a desiccator.
     Weigh  the cooled sample to  the  nearest 0.1 mg.
     Set  up a nest of sieves that will divide the coarse fraction into
     the  desired  number of subfractions.  Set up  the sieves in a graded
     series of mesh sizes, with  the  coarsest mesh on top and the finest
     mesh  on the bottom.  The  bottom sieve always should  have  a  mesh
     size of 62.5 urn (4  phi).   Place a solid  pan on the bottom of the
     stack  and a  lid on top of  the stack.   At  a minimum,  the  coarse
     fraction should  be separated into  gravel and sand fractions, using
     a sieve with a mesh  size of 2 mm (-1 phi).
     Add  the sample  to  the  uppermost sieve.   Complete transfer can be
     ensured by using a sieve brush to remove any material adhering to
     the  beaker.   The  sieve  brush can  also be used to gently break up
     aggregated sediment.
     Shake mechanically for  exactly  15  min  using  the Ro-Tap (or equiva-
     lent).  A shaker having an  automatic timer is preferable.
     After  shaking,  empty  the  contents  of each sieve onto  a glossy
     piece of paper  (e.g., wax  paper).  To empty  a sieve, invert it and
     tap  it  on  the  table several times  while  ensuring that all edges
     hit  the  table   at  the  same  time.   If the  sieve is  not  tapped
     evenly, the  meshes may be  distorted.  After  tapping the sieve,
     ensure complete  removal of the sample by brushing the back of the
     screen.  After  brushing the back of  the screen, turn  the sieve
     over  and  brush  out any particles adhering  to  the  sides  of the
     sieve or the inside of  the screen.
     Add  the  fraction  that  passed  through  the  bottom sieve  (e.g.,  4
     phi)  and was retained  by  the solid  pan  to  the silt-clay fraction
     of that sample.
     Weigh each remaining size  fraction to  the  nearest 0.1 mg.
     Sum  the weights  of all  size  fractions and compare the result with
     the   initial  weight of  the  coarse  fraction.   Losses and  inac-
     curacies should be less  than  5  percent  of the  initial  weight.
     Losses and inaccuracies tend to increase with  increasing number of
     fractions.
     Large amounts of organically derived fragments  (e.g.,  wood debris,
     grass, shells) or  any unusual material in  any size fraction should
     be noted on the  laboratory log sheet.

Silt-clay fraction -  this fraction  is  subdivided further using a  pipet
technique  that  depends upon   the differential  settling  rates of
different  particles.    Because  additions  to this  fraction  may  be made
after mechanical  sieving of  the gravel-sand fraction  (see above),  it is
recommended  that the   silt-clay  analysis  for  each sample not  be
conducted until the gravel-sand analysis  has  been completed.
     Add  10 mL  of  the dispersant  to 990  mL of  distilled water.
     Determine the weight  of  dispersant in  a  20-mL  aliquot  of this

                                 12

-------
                                   Conventional Sediment Variables
                                                    Particle Size
                                                       March 1986
mixture  by  pipeting a  20-mL  aliquot  into  each of  five  tared
beakers,  drying the  samples  to  constant  weight at  90 +_  2  C,
cooling the samples in a desiccator, weighing the cooled samples,
and  calculating the  mean weight  of dispersant  in the five
samples.  This weight multiplied by the number of 10-mL additions
of dispersant to each sample will  be subtracted from the weight of
each sediment fraction  at the  end  of the pipet analysis.
Add 10  ml  of the dispersant to each sample  suspension in the 1-L
graduated cylinders.
Mix  each  suspension  by  either  stoppering  and  inverting the
cylinder or by using the up and down motion  of  a perforated disc
plunger.
Allow the mixed  suspension to stand  for 2-3  h and check for  signs
of flocculation.  Flocculation can be recognized  by a  curdling and
rapid settling of lumps of particles or by  the presence of a  thick
soupy layer  on  the bottom of  the cylinder passing  abruptly into
clear water above.
If  flocculation  occurs,  add dispersant in 10-mL increments  until
no  noticeable   flocculation  occurs.   Record  the  volume  of
dispersant added.
When ready to conduct the  pipet analysis,  bring  the sample  volume
to  1 L  by  adding distilled water, mix the suspension thoroughly,
and place  the cylinder in a constant-temperature water bath.  If
the volume is greater or less  than 1 L, the  factor for converting
the weight  of the sediment in  each  20-mL  aliquot  to that  in the
total volume must be modified  accordingly.
After 20 sec, withdraw a 20-mL aliquot from a depth of 20 cm below
the surface of the suspension  using a pipet.  The pipet should be
marked  for  the  specified  sampling  depths  and  should be  inserted
vertically into  the  settling  cylinder  when the  aliquot is  taken.
A  suction  bulb may  be  used on  the open  end  of the  pipet  to
facilitate  sampling.   It  is  critical  that the  suspension  be
disturbed as little as possible when pipet  aliquots are taken.
Transfer the  20-mL aliquot to  a  preweighed  50-mL  beaker.   Rinse
the pipet into the beaker using 20 mL of distilled water.
Withdraw 20-mL  aliquots  at a  depth of 10  cm below the surface of
the suspension  at  the appropriate time(s) listed  in  Table 3.   A
formula  for  calculating  withdrawal  times  is  given  by Folk  (1968)
and Buchanan  (1984).   If  a withdrawal is  missed,  the suspension
can be stirred again and the missed withdrawal can be  taken  at the
appropriate  time after settling begins.    It is  not   necessary  to
withdraw the  initial 20-mL aliquot when this corrective action is
conducted.
Transfer these additional  20-mL  aliquots to  50-mL preweighed
beakers, each  time rinsing the pipet  into the   respective  beaker
using 20 mL of distilled water
Dry all aliquots to  constant weight  at 90  +  2° C.  A  drying temp-
erature less than 100° C  is used  to prevenT  boiling  and potential
loss of sample.

                            13

-------
TABLE 3. WITHDRAWAL TIMES FOR PIPET ANALYSIS AS A FUNCTION
OF PARTICLE SIZE AND WATER TEMPERATURE3 ••
Diameter
Finer
than
(phi)
4.0
5.0
6.0
7.0
8.QC
9.0
10.0
Diameter
Finer Withdrawal
than Depth
(urn) (cm) 18° C
62.5
31.2
15.6
7.8
3.9
1.95
0.98
20
10
10
10
10
10
10
20s
2mOs
8mOs
31m59s
2h8m
8h32m
34h6m
Elapsed Time for Withdrawal of Sample in
Hours (h) , Minutes (m), and Seconds (s)
19° C 20° C 21° C 22° C 23° C 24° C
20s
Im57s
7m48s
31ml Is
2h5m
8hl8m
33hl6m
20s
Im54s
7m36s
30m26s
2h2m
8h6m
32h28m
20s
Im51s
7m25s
29m41s
Ih59m
7h56m
31h40m
20s
Im49s
7ml 5s
28m59s
Ih56m
7h44m
3 Oh 56m
20s
Im46s
7m5s
28ml8s
Ih53m
7h32m
30hl2m
20s
Im44s
6m55s
27m39s
Ih51m
7h22m
29h30m
a Modified from Plumb (1981).
b It is critical that temperature be held constant during the pipet analysis.
c Breakpoint between silt and clay.

-------
                                             Conventional Sediment Variables
                                                              Particle Size
                                                                 March 1986
          Cool dried samples  to room temperature  in a desiccator.
          Weigh cooled samples to the nearest 0.1 mg.
•    Calculations
     -    The  total  weight of  a phi-size  interval  in the  1-L graduated
          cylinder is determined  as  follows:

                    Phi weight  (g dry weight)  =  50[(A-C)-(B-C)]

          Where:

               A = weight (g)  of  residue in  a  20-ml  aliquot for a
                   given phi-size boundary
               B = weight (g)  residue in a 20-mL aliquot  for the  next
                   larger phi-size boundary
               C = mean weight  (g) of dispersant in  a  20-mL aliquot.

QA/QC Procedures

     It  is critical  that each sample be homogenized thoroughly  in the
laboratory before  a  subsample is taken  for analysis.   Laboratory
homogenization should  be  conducted  even if  samples  were homogenized in the
field.

     After  dry-sieving a  sample,  all  material  must be removed  from the
sieve.  This can be accomplished  by  tapping  the  rim  of the sieve  evenly on a
hard surface and by brushing the  screen.

     The  total  amount  of fine-grained material used  for  pipet analysis
should be  5-25  g.   If more material  is  used,  particles  may  interfere with
each  other  during  settling  and the  possibility  of  flocculation  may be
enhanced.   If  less  material  is   used,  the  experimental error  in weighing
becomes large relative to the sample size.

     Before pipet  extractions  can be made, the sample  must  be  homogenized
thoroughly within  the  settling  cylinder.   Once the pipet analysis begins,
the settling cylinders  must not  be   disturbed,  as this  will  alter particle
settling velocities.  Care  must  be taken to disturb the  sample as  little as
possible when pipet extractions are  made.

     After  a  pipet  extract has   been  transferred to a  drying  beaker, any
sample adhering  to the inside of the pipet must be  removed.  This  can be
accomplished by  drawing 20 mL  of distilled water into the pipet and adding
this rinse water to the drying  beaker.

     Dried  samples should  be  cooled in a  desiccator and  held  there  until
they are weighed.  If a desiccator is not used,  the  sediment will accumulate
ambient  moisture and  the sample weight will  be overestimated.   A color-
indicating desiccant  is recommended  so  that spent desiccant can  be detected

                                      15

-------
                                             Conventional  Sediment Variables
                                                              Particle Size
                                                                 March 1986
easily.  Also,  the  seal  on the  desiccator  should  be checked periodically,
and, if necessary, the ground glass rims should  be greased  or the  "0" rings
should be replaced.

     It is recommended that triplicate analyses  be  conducted on  one of every
20  samples,  or on  one sample   per  batch  if  less   than  20  samples  are
analyzed.  It is also recommended that the  analytical balance,  drying oven,
and temperature bath  be  inspected daily and  calibrated  at  least  once per
week.

DATA REPORTING REQUIREMENTS

     The weight of each sediment  fraction should be  reported to the nearest
0.0001  g  dry weight.   The laboratory  should  report  the  results  of  all
samples analyzed (including QA replicates)  and  should note any problems  that
may have influenced  data quality.
                                      16

-------
                                            Conventional Sediment Variables
                                                               Total  Solids
                                                                 March  1986
                               TOTAL SOLIDS
USE AND LIMITATIONS

     Total solids are the organic and inorganic materials remaining after a
sample has been dried completely.  This variable is commonly used to convert
sediment  concentrations of  substances  from a  wet-weight  to  a  dry-weight
basis.  It typically is  measured  in conjunction with other variables.

     Total solids  values are operationally  defined,  because results depend
on drying  temperatures.   For example, temperature-dependent weight  losses
occur  from  volatilization  of organic matter,  mechanically  occluded water,
water  of  crystallization,  and  gases  from  heat-induced  chemical
decomposition.   By  contrast,  weight  gains  may result  from  oxidation
processes.  To provide data that are comparable among different studies, it
is therefore critical that  drying temperatures  be standardized.

FIELD PROCEDURES

Collection

     Samples can  be  collected  in glass  or plastic  containers.   Samples can
be collected in the same containers as  samples  for  other  variables, if total
solids  is  to  be  measured  in conjunction with  those  variables.   A minimum
sample size of 50  g  is  recommended.   If  unrepresentative material  is  to be
removed from  the sample,  it  should  be removed in the   field  under the
supervision of the chief scientist and  noted  on the field log sheet.

Processing

     Samples should  be  stored frozen and can  be held  for  up  to 6 mo under
that condition.

LABORATORY PROCEDURES

Analytical Procedures

•    Equipment
     -Muffle furnace
               550° C capacity
          Drying oven
          Desiccator
                                      17

-------
                                            Conventional Sediment Variables
                                                               Total Solids
                                                                 March 1986
          Analytical  balance
               0.01  g accuracy
          100-mL evaporating  dishes
               Porcelain,  platinum,  or Vycor.

•    Equipment preparation
     -Ignite clean evaporating dishes  at 550 +_ 10° C for 1 h in a muffle
          furnace to remove any  remaining  organic  material.
          Cool ignited dishes to room temperature  in a desiccator.
          Weigh each  cooled  dish to the  nearest 0.01  g  and store  in the
          desiccator.

•    Sample preparation
     -Allow frozen sediment  samples  to warm to room temperature
          Homogenize each sample mechanically,  incorporating  any overlying
          water.
          Transfer  a  representative subsample  (approximately  25 g)  to a
          preweighed evaporation dish.
          Weigh the undried  sample to the  nearest  0.01 g.

•    Analytical procedures
     -    Dry the sample to  constant weight at 103 ^ 2° C.
          Cool the dried sample  to room  temperature  in a desiccator.
          Weigh the cooled sample to the nearest 0.01 g.

•    Calculations
     -Total solids content  is determined as follows:
                    Percent sol Ids  -
          Where:
               A = weight (g)  of dish and dry  sample  residue
               B = weight (g)  of dish
               C = weight (g)  of dish and wet  sample.

QA/QC Procedures

     It  is  critical  that each sample  be thoroughly  homogenized  in the
laboratory  before  a subsample  is  taken  for analysis.    Laboratory
homogenization should  be  conducted  even  if samples  were homogenized in the
field.

     Evaporating  dishes  must  be  ignited at  550° C  before  being used for
total solids analysis.  This step ensures that dishes are free  from organic
contaminants.

     Dried  samples  should be cooled in  a  desiccator and  held there  until
they are weighed.  If a desiccator is not used,  the  sediment will  accumulate

                                      18

-------
                                             Conventional  Sediment  Variables
                                                                Total  Solids
                                                                 March  1986
ambient moisture  and the  sample weight  will  be  overestimated.   A  color-
indicating desiccant is recommended so that  spent  desiccant  can  be  detected
easily.  Also,  the seal on  the desiccator  should be checked periodically
and, if necessary, the  ground glass rims  should  be greased or the "0" rings
should be replaced.

     It is recommended that triplicate analyses be conducted on one  of every
20 samples or on one sample per batch if  less  than 20  samples are analyzed.
It  is  also recommended  that  the analytical  balance and  drying  oven  be
inspected daily and calibrated at least  once per week.

DATA REPORTING REQUIREMENTS

     Total solids  should  be reported as  a percentage of  the wet weight  of
the  sample  to  the nearest 0.1  unit.  The  laboratory should  report the
results of  all  samples analyzed  (including  QA replicates) and  should  note
any problems that may have influenced  sample quality.
                                      19

-------
                                             Conventional  Sediment  Variables
                                                 Total  Volatile Solids  (TVS)
                                                                 March 1986
                        TOTAL VOLATILE SOLIDS (TVS)
USE AND LIMITATIONS
     Total volatile solids  represent  the fraction of total solids  that  are
lost on ignition at a  higher  temperature than that used to determine  total
solids.  Total volatile solids  is used  as  a  crude estimate of the amount of
organic matter in the total  solids.

     Total volatile solids  is operationally  defined by  the ignition temper-
ature.  Total volatile solids content does not always represent  the organic
content of a sample because some  organic material  may be lost at the drying
temperature and some inorganic material  (e.g., carbonates,  chlorides) may be
lost at the  ignition temperature.   Because of the temperature dependence of
total  volatile  solids, valid  interstudy comparisons  require  the  use  of
standardized drying and ignition temperatures.

FIELD PROCEDURES

Collection

     Samples can  be collected  in glass or  plastic  containers.    A minimum
sample size of 50 g  is recommended.  If unrepresentative material  is  to be
removed  from  the  sample,  it should  be removed  in  the  field  under  the
supervision of the chief scientist and noted  on the field log  sheet.

Processing

     Samples should  be stored frozen and can be  held  for up to  6 mo under
that condition.

LABORATORY PROCEDURES

Analytical Procedures

•    Equipment
     -Muffle furnace
               550° C  capacity
          Drying oven
          Desiccator
          Analytical balance
               0.01 g  accuracy
          100-mL evaporating dishes
               Porcelain,  platinum, or Vycor.


                                      20

-------
                                             Conventional Sediment Variables
                                                Total Volatile Solids (TVS)
                                                                 March 1986
•    Equipment preparation
     -    Ignite clean evaporating dishes  at  550 +_ 10° C  for 1 h  in a muffle
          furnace to remove any  remaining  organic  material.
          Cool ignited dishes to room temperature  in  a desiccator.
          Weigh  each  cooled dish  to  the  nearest  0.01  g and store  in the
          desiccator.

t    Sample preparation
     -    Allow frozen sediment  samples to warm to room temperature.
          Homogenize each  sample  mechanically, incorporating  any overlying
          water.
          Transfer  a  representative  subsample  (approximately  25 g)  to a
          preweighed evaporating dish.

•    Analytical procedures
     -    Dry the sample to constant weight at  103 +_ 2° C.
          Cool the dried sample  to room temperature in a  desiccator.
          Weigh the cooled sample to the nearest 0.01 g.
          Ignite the  sample  at  550 +  10° C  to constant weight.  Make sure
          that  the  samples  do  not flare up when  placed  in the oven,  as
          sediment may be lost from the crucibles.   If sample flashing is a
          problem, it is  recommended that the  muffle furnace be  cooler than
          550° C  when  samples are  placed inside,  and  that the  temperature
          gradually be increased to 550° C.
          Weigh each cooled sample to the  nearest  0.01 g.

•    Calculations
     :    TVS content is determined as follows:

                         Perc(int TVS . i
          Where:

               A = weight (g) of dish and dry sample residue
               B = weight (g) of evaporation dish
               C = weight (g) of dish and ignition residue.

QA/QC Procedures

     It  is  critical  that  each  sample  be thoroughly  homogenized  in  the
laboratory  before  a  subsample  is  taken  for   analysis.   Laboratory
homogenization  should  be  conducted even if samples were homogenized in  the
field.

     Evaporating  dishes (or  crucibles)  must be  ignited at  550°  C before
being used for  total  volatile solids analysis.   This step ensures that  the
dishes are free from volatile contaminants.


                                      21

-------
                                             Conventional Sediment Variables
                                                Total Volatile Solids (TVS)
                                                                 March 1986
     Dried and combusted samples should be  cooled  in  a desiccator and held
there until  they are weighed.   If a desiccator is not  used, the sediment
will  accumulate ambient  moisture  and  the  sample  weight  will  be
overestimated.   A  color-indicating desiccant is recommended  so that spent
desiccant can be detected easily.  Also,  the seal  on  the desiccator should
be checked periodically  and,  if necessary,  the  ground glass  rims  should be
greased or the "0" rings should be  replaced.

     It is recommended that  triplicate analyses  be  conducted on one of every
20 samples or on one sample  per batch if less than 20 samples are  analyzed.
It is also recommended that the analytical balance, drying oven, and muffle
furnace be inspected daily and calibrated  at  least  once per week.

DATA REPORTING REQUIREMENTS

     Total volatile  solids  should be  reported  as a  percentage  of the dry
weight of the uncombusted sample  to the  nearest 0.1  unit.   The laboratory
should report the results of  all samples  analyzed  (including  QA  replicates)
and should note any problems that  may have influenced  data quality.
                                      22

-------
                                             Conventional Sediment Variables
                                                 Total Organic Carbon (TOC)
                                                                 March 1986
                        TOTAL ORGANIC CARBON (TOC)
USE AND LIMITATIONS
     Total organic carbon  is  a  measure of the total amount of nonvolatile,
volatile,  partially  volatile,  and  particulate  organic compounds  in  a
sample.  Total organic  carbon  is  independent  of the oxidation state of the
organic  compounds  and  is  not  a  measure of  the  organically  bound  and
inorganic elements  that  can  contribute  to  the  biochemical  and chemical
oxygen demand tests.

     Because inorganic carbon  (e.g.,  carbonates,  bicarbonates,  free C02)
will interfere with  total  organic  carbon  determinations,  samples  should be
treated to remove inorganic carbon  before  being  analyzed.

FIELD PROCEDURES

Collection

     Samples  can  be  collected  in  glass  or plastic  containers.   A  minimum
sample size  of 25 g  is  recommended.   If  unrepresentative  material is to be
removed  from the  sample,  it should  be removed  in  the field  under the
supervision of the chief scientist  and  noted on  the field  log  sheet.

Processing

     Samples  should  be  stored frozen and can be  held  for up  to 6 mo under
that condition.  Excessive temperatures should not be used to  thaw  samples.

LABORATORY PROCEDURES

Analytical Procedures

•    Equipment
     -Induction furnace
               e.g., Leco WR-12, Dohrmann  DC-50, Coleman CH analyzer,
               Perk in Elmer 240 elemental  analyzer, Carlo-Erba 1106
          Analytical balance
               0.1 mg accuracy
          Desiccator
          Combustion boats
          10 percent hydrochloric acid (HC1)
          Cupric oxide fines (or equivalent material)
          Benzoic acid or other carbon source  as a standard.
                                      23

-------
                                            Conventional Sediment Variables
                                                 Total Organic Carbon (TOC)
                                                                 March 1986

•    Equipment preparation
     -Clean combustion  boats  by  placing them in the induction furnace at
          950°  C.   After  being cleaned, combustion  boats  should  not  be
          touched with  bare hands.
          Cool boats to room  temperature  in a desiccator.
          Weigh each boat to  the  nearest  0.1 mg.

§    Sample preparation
     •Allow frozen  samples  to warm  to room temperature.
          Homogenize each  sample mechanically,  incorporating  any overlying
          water.
          Transfer  a representative  aliquot (5-10 g) to a clean container.

•    Analytical procedures
     -    Dry  samples  to constant weight  at 70  +_ 2°  C.   The  drying
          temperature is relatively  low to minimize loss of volatile organic
          compounds.
          Cool dried samples  to room temperature in a desiccator.
          Grind sample  using  a  mortar and pestle to break up aggregates.
          Transfer  a  representative  aliquot  (0.2-0.5  g)  to a  clean,
          preweighed combustion boat.
          Determine sample  weight to the  nearest 0.1 mg.
          Add  several  drops  of HC1  to the dried  sample to  remove car-
          bonates.   Wait until  the  effervescing is completed and  add more
          acid.  Continue  this process  until  the  incremental  addition  of
          acid causes no further effervescence.  Do not add too much acid at
          one  time  as this  may cause  loss  of  sample  due  to frothing.
          Exposure   of  small  samples (i.e.,  1-10 mg)  having  less  than  50
          percent carbonate to an HC1 atmosphere for 24-48 h has been shown
          to be an effective means of  removing carbonates (Hedges and Stern
          1984).  If this  method is used for sample  sizes  greater than 10
          mg, its effectiveness should  be demonstrated by the user.
          Dry the HCl-treated sample to constant weight at 70 ^2° C.
          Cool to room temperature in a desiccator.
          Add  previously  ashed  cupric oxide  fines or  equivalent material
          (e.g., alumina oxide) to the  sample  in the combustion boat.
          Combust  the  sample in  an  induction  furnace  at  a  minimum
          temperature of 950  ^10° C.

t    Calculations
     -If  an ascarite-filled tube  is used  to  capture COg,  the  carbon
          content of the sample can  be  calculated as follows:

                       Percent carbon . A(Q.27|9)(100)
                                      24

-------
                                            Conventional  Sediment  Variables
                                                 Total  Organic Carbon  (TOC)
                                                                March  1986
          Where:
               A =   the weight  (g)  of  C02 determined  by  weighing the
                    ascarite tube before and after combustion
               B =   dry weight  (g)  of the  unacidified sample  in the
                    combustion  boat
          0.2729 =   the ratio  of  the  molecular  weight  of  carbon  to the
                    molecular weight of carbon dioxide

          A silica  gel  trap  should be  placed  before the ascarite tube  to
          catch  any  moisture  driven  off during  sample  combustion.
          Additional silica gel  should be  placed  at the  exit  end  of the
          ascarite  tube to trap any water that might be  formed  by reaction
          of the trapped C02 with the NaOH in the ascarite.
          If an elemental  analyzer  is used,  the  amount of  CQ.2 will  be
          measured   by  a thermal conductivity  detector.   The  instrument
          should be calibrated  daily using an empty  boat  blank  as the zero
          point and at  least two standards.   Standards  should  bracket the
          expected  range of carbon  concentrations in the samples.

Q.A/QC Procedures

     It  is critical   that  each  sample be  thoroughly  homogenized  in the
laboratory before a  subsample  is  taken  for  analysis.   Laboratory
homogenization  should be conducted even  if  samples were  homogenized  in the
field.

     Dried  samples  should  be   cooled  in  a  desiccator and held  there  until
they are weighed.  If  a desiccator  is not used, the sediment will accumulate
ambient  moisture  and  the  sample weight will  be  overestimated.  A  color-
indicating desiccant is recommended so that  spent  desiccant  can  be detected
easily.  Also,  the seal on the desiccator  should be  checked  periodically
and, if necessary,  the ground glass rims should  be  greased or the "0"  rings
should be replaced.

     It is recommended that triplicate  analyses be conducted on one of every
20  samples,  or  on  one sample per  batch  if  less  than  20 samples are
analyzed.  A method blank  should be analyzed at the same frequency  as the
triplicate analyses.  The analytical balance should be  inspected  daily and
calibrated at least once per week.   The carbon analyzer should be calibrated
daily with freshly  prepared standards.   A standard reference material  should
be analyzed at  least once  for each  major survey.

DATA REPORTING  REQUIREMENTS

     Total  organic  carbon should  be  reported as  a  percentage of the dry
weight of  the  unacidified sample  to the nearest 0.1 unit.   The laboratory
should report  the  results  of   all  samples  (including QA  replicates,  method
                                     25

-------
                                             Conventional Sediment Variables
                                                  Total  Organic Carbon (TOC)
                                                                  March 1986

blanks, and  standard  reference measurements) and  should note  any  problems
that may have influenced sample quality.  The laboratory should also provide
a summary  of the  calibration  procedure and  results  (e.g.,  range  covered,
regression equation, coefficient of determination).
                                      26

-------
                                             Conventional  Sediment  Variables
                                          Oil  and  Grease  (Freon-Extractable)
                                                                 March 1986
                    OIL AND GREASE  (FREON-EXTRACTABLE)
USE AND LIMITATIONS
     Oil and  grease  tests  measure  all  material recovered  as a  substance
soluble in  a  nonpolar solvent  (e.g.,  Freon)  under acidic  conditions.   Oil
and grease  includes  such compounds  as hydrocarbons,  vegetable oils,  animal
fats, waxes, soaps, greases, and related industrial  compounds.

     In addition to oil  and  grease,  the  solvent may dissolve other kinds of
substances, such  as  sulfur compounds, organic  dyes,  and  chlorophyll.   Oil
and grease  is therefore  operationally defined  by  the kind  of  solvent and
analytical   methods  used.   Standardized  procedures  are essential  for valid
interstudy  comparisons.   Because oil  and  grease values  obtained  using the
gravimetric technique sometimes do not correlate with values obtained using
the IR  technique,  it  is recommended that  a  single  technique be used within
each study  (at a minimum) and among different studies (if possible).

FIELD PROCEDURES

Collection

     Samples  should  be collected only in  glass containers having TFE-lined
lids.   Although  aluminum-lined lids can  be  used,  seawater  eventually will
corrode the aluminum.  Before being used, containers and lids should  first
be  washed   with  a warm  aqueous detergent mixture  and  then, in  sequence,
thoroughly  rinsed  with  hot tap water, rinsed  at least twice  with distilled
water,  rinsed  once  with Freon (i.e.,  l,l,2-trichloro-l,2,2-trifluoroethane,
and  dried   at  105  +_  2°  C for  30  min.  A minimum  sample size of 100 g  is
recommended.   If unrepresentative material is to be removed  from the  sample,
it  should   be  removed  in  the field under the supervision  of  the chief
scientist and  noted on  the  field log sheet.

Processing

      If samples  cannot be analyzed  within 24  h, they can be  preserved with
approximately  1 mL  of  concentrated hydrochloric  acid  (HC1) per 80  g  of
sample.   Acid-preserved  samples  should  be stored  at 4°  C, and can  be  held
for up  to 28  days  in  that condition.  Although U.S.  EPA has  not established
a  recommeded maximum holding  time for oil  and grease in  sediments,  28 days
is  consistent with  the recommended holding  time  for  acid-preserved  water
samples.    Samples can  also be preserved by freezing at  -20° C, and  can  be
held  under  that  condition for up  to  6 mo.  Samples must be  kept field-moist
during  storage because they may lose  apparent  oil  and grease as a result  of
drying.
                                      27

-------
                                             Conventional  Sediment Variables
                                          Oil  and  Grease  (Freon-Extractable)
                                                                 March  1986

LABORATORY PROCEDURES

Analytical Procedures

•    Equipment
     -Infrared spectrophotometer
               IR technique only
          Analytical balance
               Gravimetric technique only, 0.1 mg  accuracy
          Extraction apparatus, Soxhlet
          Vacuum pump or other source of vacuum
          Extraction thimble, paper
          Concentrated hydrochloric acid (HC1) or  concentrated  sulfuric  acid
          (H2S04)
          Magnesium sulfate monohydrate
               Prepare  MgSO^HgO  by  drying a  thin  layer of MgS04«7H20
               overnight in an oven at  103° C
          Freon  (l,l,2-trichloro-l,2,2,-trifluoroethane), boiling point  47°
          C
               The  solvent  should  leave  no  measurable  residue- on  evapor-
               ation; redistill if necessary
          Grease-free cotton
               Extract nonabsorbent cotton using Freon
          Oil reference standard
               If the  identity  of oil  and grease in  a sample  is unknown,  a
               mixture of 15.0 ml n-hexadecane, 15.0  ml  isooctane, and  10.0
               ml chlorobenzene should be used as  the standard.   This is the
               same reference oil used for water  samples  in  U.S. EPA Method
               413.2 (U.S. EPA 1983).  If the identity of oil  and grease is
               known, the standard can  be comprised  of the same substance as
               that in the sample.

•    Sample preparation
     •Allow samples to warm to room temperature.
          Homogenize each  sample mechanically, incorporating any overlying
          water.
          Remove  a  representative aliquot (approximately 25 g)  and  analyze
          it for total solids content.
          Remove  a  representative aliquot  (approximately  20  g)  and weigh it
          to the nearest 0.1 mg.
          Transfer the weighed aliquot to a 150-mL beaker for oil and grease
          analysis.

•    Oil and grease extraction
     -    Acidify  the  sample to  pH  =  2  using concentrated  HC1   or  con-
          centrated H2S04.
          Add 25 g MgSO^HgO to the acidified sediment sample.   Stir  to  make
          a uniformly smooth paste that is spread  on the  beaker wall. Allow
          to stand 15-30 min until solidified.


                                      28

-------
                                        Conventional  Sediment Variables
                                     Oil  and Grease  (Freon-Extractable)
                                                            March 1986

     Following  solidification,  remove the  solids  and grind  in a
     porcelain mortar.   The use  of a  desiccated,  uniformly  ground
     sample improves the efficiency  of  the  extraction process.
     Add the ground  sample  to  a paper  extraction thimble.  The beaker
     and mortar  should be wiped with a  small piece  of filter paper that
     has  been soaked  in Freon.  Add  the  filter  paper  to  the paper
     thimble.
     Fill the thimble  with  glass wool  or  small  glass beads.   Extract
     the prepared sample using Freon in  a  Soxhlet  apparatus  at a rate
     of 20 cycles/h for 4 h.   Stir  at  least twice  during  extraction  to
     prevent channeling.  If  the final  extract  is  turbid, filter  the
     sample through  grease-free  cotton into  a clean flask.  Rinse  the
     initial sample  container  and  the  cotton with  Freon  and  add  the
     washing to  the filtered sample.
     Oil  and grease  concentration  of  the extract  can  be determined
     using  either  the  infrared  spectrophotometry  or  the gravimetric
     method.

Infrared spectrophotometry method
-Quantitatively transfer the sediment  extract to a convenient size
     volumetric  flask and dilute to  volume  with Freon.
     Prepare calibration standards using the  reference  oil.
     Transfer required  amounts of  the  reference material into 100-mL
     volumetric   flasks  using  microliter  pipettes.   Dilute to  volume
     with Freon.
     The most appropriate pathlength for the  quartz cells  to be  used in
     the spectrophotometric  determination is  determined by the expected
     sample concentration.  The following information is presented as a
     guide for selecting cell  length:
          Pathlength. cm                Expected  Range, mg
                T~~^                          4  -  40
                 5                           0.5-8
                10                           0.1  -  4
     Based on observed ranges of oil and grease in  sediments,  it may be
     necessary  to  dilute  the  sample  extracts   to  the working ranges
     indicated above.
     Scan the standards and samples from  3,200  to  2,700 cm'1  using a
     recording infrared spectrophotometer.   Freon should be used in  the
     reference beam of  a dual  beam  instrument or to zero  a single beam
     instrument.   The  absorbance  of the 2930-cm"1  peak should  be used
     to construct a standard curve.
     Prepare a standard curve  by plotting measured  absorbance  vs.  oil
     and grease  concentration of the standards.   Compare the absorbance
     of  the  Freon  extract to  the standard curve to determine  the  oil
     and grease  concentration.
     Calculate oil and grease concentration as follows:

          Concentration (mg/kg  dry weight)  =
                                 29

-------
                                            Conventional Sediment Variables
                                         Oil and Grease (Freon-Extractable)
                                                                 March 1986

          Where:
                  X = concentration  of  oil and grease in the Freon
                      extract,  mg/L
                  V = volume  of Freon extract, L
                  g = wet  weight of  sediment extracted, g
                % S = percent total  solids in the sediment sample
                      (expressed as  a decimal fraction).

•    Gravimetric method
     •Quantitatively transfer the sediment extract to a tared distilling
          flask.   Rinse  the  extract container  with Freon  and  add  to  the
          distilling flask.
          Distill the Freon  from the extraction flasks using a water bath at
          70° C.
          After the solvent  has  been evaporated, place  the  flask on a warm
          steam bath for 15 min and draw air through the  flask  by  means of
          an applied vacuum for the  final 1 min.
          Cool the sample  in  a  desiccator for 30 min.
          Weigh the cooled sample to the nearest 0.1 mg.
          Calculate oil and  grease concentration as  follows:


          Concentration (mg/kg  dry weight) =  (A


          Where:

               A =  weight (mg)  of  tared  flask  and oil  and grease residue
               B = weight  (mg)  of tared flask
               g = wet weight (g) of sediment extracted
              %S =  percent  total  solids  in the sediment  sample (expressed
                   as a decimal fraction)
            1000 = factor  for converting mg/g to mg/kg.

QA/QC Procedures

     Because the  results  of  an  oil  and  grease analysis  are extremely
sensitive to  the methods  used, comparable  results   can be  obtained  only by
strict adherence to all methodological  details.

     It  is  critical  that  each sample be  thoroughly  homogenized   in  the
laboratory  before  a  subsample  is taken  for  analysis.    Laboratory
homogenization should  be  conducted  even if  samples  were homogenized in the
field.

     It is recommended that triplicate  analyses  be conducted on  one  of every
20  samples,  or  on  one sample  per batch  if  less than 20  samples  are
analyzed.  Also, a method blank should  be analyzed for each  batch of samples
extracted.
                                      30

-------
                                             Conventional  Sediment Variables
                                          Oil and Grease (Freon-Extractable)
                                                                  March  1986
DATA REPORTING REQUIREMENTS
     Oil and grease concentrations should be reported as mg/kg dry weight to
no more  than  three significant figures.   The  laboratory should  report  the
results of  all  samples  (including QA  replicates  and  method blanks)  and
should  note  any  problems  that may  have  influenced  sample quality.   The
laboratory should  also provide a description of  the calibration  procedures
and standards  used to determine  oil  and grease concentrations by  infrared
spectrophotometry.
                                      31

-------
                                             Conventional  Sediment  Variables
                                                              Total  Sulfides
                                                                 March  1986
                              TOTAL SULFIDES
USE AND LIMITATIONS

     Total sulfldes represent  the  amount of acid-soluble H2S, HS",  and  S2~
in a sample.   Sulfides  are measured because they may  be toxic and  because
they may  create unaesthetic  conditions.   This method  cannot  be used if  a
measure of  only water-soluble  sulfides  is desired.   A  measure  of  water-
soluble sulfides might  be  desired  if an estimate of  biologically  available
sulfides is needed.

     Sulfides  are difficult  to  sample because  some may  be lost  through
volatilization  and/or gas  stripping  and  some may be  lost  through  oxidation
by dissolved oxygen.

FIELD PROCEDURES

Collection

     Samples  can  be collected  in  glass or  plastic containers.   A  minimum
sample size of  50  g  is  recommended.   If unrepresentative material is  to be
removed  from  the sample,  it should  be  removed  in  the  field under  the
supervision of  the chief scientist and noted on the field log sheet.

Processing

     Samples for total sulfides should be preserved in the field  by adding 2
N  zinc  acetate  solution   (approximately 5 mL  per  30  g  of sediment)  and
swirling  the  mixture.   Samples should  be stored  in  the  dark at 4°  C  and
analyzed as soon as possible.   It  is  critical  that  air contact with  samples
be  minimized  and  samples  be kept moist,  to minimize  oxidation.   Although
U.S. EPA has not established a recommended maximum holding time for sulfides
in  sediments,  a maximum holding time of  7  days would be consistent with  the
holding time recommended for sulfldes in preserved water samples.

LABORATORY PROCEDURES

Analytical Procedures

•    Equipment
     -Distillation apparatus, all glass
                For large  samples,  a  suitable assembly  consists  of  a  1-L
                pyrex distilling flask with  Graham condenser  as used  for  the
                analysis  of phenols.   A  section  of glass tubing  should be


                                      32

-------
                                   Conventional  Sediment  Variables
                                                   Total Sulfides
                                                       March  1986

     connected to the tip of the condenser so that it  reaches  the
     bottom of the collection tube.
Distillate collection tubes
     Short-form Messier tubes,  graduated at 50  and 100  ml.
Spectrophotometer
     For  use  at  650  nm and  providing a  light path  of 1  in  or
     greater.
Nitrogen, water-pumped
Zinc acetate,  2 N
     Dissolve  2215 g  of  ln(Cz^Qz)Z'2^ZQ  in distilled water  and
     dilute to 1 L.
Zinc acetate,  0.2 N
     To 100 ml of 2~^Zn(C2H302)2'2H20 add several drops  of  acetic
     acid and  dilute to 1 L.
Sulfuric acid solution, 1:1
     Carefully  add  500 ml  of concentrated H2S04 to  500  ml  of
     distilled water  in  a 1-L flask.   Mix continuously and  cool
     under running water while combining  reagents.   Cool solution
     before using.
Dilute sulfuric acid solution,  approximately 0.1 N^
     Dilute 5 mL of 1:1 ^$04 to 1 L with distilled water.
Stock amine solution
     Dissolve  2.7 g  of  N,  N-dimethyl-p-phenylenediamine  sulfate
     and dilute  to 100 ml with  1:1 H2S04  solution.   This solution
     is stable for approximately 1 wk.
Working amine solution
     Dilute 2 ml  of  stock  amine solution  to 100 ml with 1:1 H2S04
     solution.  Prepare fresh daily.
Ferric chloride  solution
     Dissolve  100 g  of  FeCl3«6H20  in  hot  distilled  water  and
     dilute to 100 ml.  Cool before use.
Standard potassium biniodate solution, 0.025 N^
     Accurately  weigh  out 0.8124  g  KH(I03)2  and  dissolve  in
     distilled water.  Dilute to 1 L.
Standard sodium  thiosulfate titrant, 0.025 N_
     Dissolve  6.205  g Na2$203'5H20 in  distilled  water and  dilute
     to 1 L.   Preserve with  5  ml  chloroform.   Standardize  against
     standard potassium biniodate  using starch  as an  indicator.
Potassium iodide  solution
     Dissolve 5  g of KI in distilled water and  dilute to 100 ml.
Treated hydrochloric acid
     Place  one  or  two strips  of aluminum  in  a small  beaker of
     concentrated  HC1.   Following  the subsequent reaction,  the
     acid is poured off and  is ready to use.
Oxygen-free dilution water
     Pass nitrogen  gas  through a  sufficient quantity of distilled
     water  for  dilution  requirements.   A  minimum of 10  min  is
     required  to displace oxygen in the water.
Sodium sulfide,  reagent, crystal.
                            33

-------
                                             Conventional Sediment Variables
                                                             Total Sulfides
                                                                 March 1986

t    Standards preparation
     :Prepare 0.01 N_ sulfide solution as follows:  weigh approximately
          1.2 g of large  crystal  Na2S-9H20.   Wash  the  crystals several times
          with distilled  water.   Discard the washings  and add  the washed
          crystals to 975 ml of  nitrogen-saturated distilled water.  Dilute
          to 1 L.
          Pipet 20 ml of stock  sulfide solution into 100 ml of oxygen-free
          water.   Add 5 ml  of  KI  solution,  20 ml  of 0.025  N KH(I03)2
          solution,  and  10 ml of 0.1 N_ H2S04.   Titrate with 0.025~^ Na2S203
          solution using  starch  as an  endpoint indicator.  Carry  a blank
          through the procedure  and  calculate  the amount of reacted iodine
          from the  difference  between  the  blank  and  standard titrations.
          Because 1 ml of 0.025 N_ KH  (103)2 is  equivalent to  0.400 mg of
          sulfide ion,  calculate the  sulfide  concentration   in  the stock
          solution.
          Calculate  the volume of  stock  solution  that  contains  0.2 mg
          sulfide and  add this  amount to  900  ml  of oxygen-free  water.
          Dilute to 1 L.  This  is  the  working  standard  containing  2 ug
          S/mL.   Sulfide  solutions are  extremely unstable  and must  be
          prepared fresh  and  used immediately.   Stability  is increased by
          using nitrogen-saturated water  for dilution.
          Prepare a  standard  curve  by dilution of the  working  sulfide
          solution.-  Pipet 20  ml 0.2 N_  Zn(C2H302)2  into a series of 50-mL
          Nessler tubes.    Add  the required  amounts  of sulfide solution to
          each Nessler tube,  taking care to pipet the  solution  below  the
          Zn(C2H302)2  level.   Dilute to 50 ml with oxygen-free water.
          Equilibrate  the temperature  of  the standards to  23-25° C using a
          water  bath while  the colorimetric reagents are  added.   Add  2 ml
          dilute amine-sulfuric acid solution to the standard, mix, and  add
          0.25 ml (5 drops) FeCl3 solution. Mix the solution and allow 10
          min for color development.   Measure the  absorbance at 650 nm.

t    Sample preparation
     :Allow samples  to warm to room temperature.
          Homogenize each  sample mechanically,  incorporating  any overlying
          water.
          Remove  a  representative  aliquot  (approximately 25 g) and  analyze
          for total  solids content.
          Remove a representative  aliquot for total sulfides analysis.   The
          aliquot should  not contain  more than  50  ug  of sulfide.
          Weigh the aliquot to the nearest 0.1  mg.

•    Distillation
     -Set  up the distillation  apparatus.   The transfer  tube from  the
          condenser should reach  the  bottom of the distillated  collection
          tube.  The condenser  should  be attached so that it can be easily
          moved up or down when diluting the distillate or  adding  reagents.
          Pipet  20 ml  of 0.2  N^ Zn(C2H302)2 into a 100-mL Nessler tube  and
          lower  the condenser so that  the transfer tubing  reaches below  the


                                      34

-------
                                            Conventional Sediment Variables
                                                             Total Sulfides
                                                                 March 1986

          level  of the liquid.   Attach  a distilling flask and pass nitrogen
          gas  through  the  system for at  least 10 min.
          Transfer the  sample  aliquot   to  the  distillation  flask.   Bubble
          nitrogen gas through  the sample to remove any oxygen dissolved in
          the  sample.   A small   amount of  sulfide  may  be driven over by the
          gas, so be  sure  that the only  exit  is  through  the  zinc  acetate
          solution in  the  collecting tube.
          Discontinue  nitrogen evolution  and  add rapidly  several  boiling
          stones,  two  drops of  methyl orange indicator,  and  enough  treated
          HC1  to change the color from orange to red.  Stopper as quickly as
          possible and  heat slowly.   The  slower  the  heating  rate,  the
          greater the  contact  time between the  evolved H2S and Zn(C2H302)2
          and  the less chance of sulfide loss.
          Distill  the  solution  until approximately 20  ml of distillate has
          been collected (roughly 5-8 min  after the solution  commences  to
          boil).   Turn off  heat  and remove the  stopper in the distillation
          flask  to  keep the distillate  from  being  sucked back  up  the
          condenser.   Raise the  transfer  tube  above the 50-mL mark  on the
          collection container  and dilute the solution to 50 ml.
          Place  the distillates  in a  water bath  at  23-25°  c.   Add  2  mL
          dilute  amine  solution and   mix.  Add  0.25  mL (5  drops)  FeCl3
          solution and mix.  Allow 10  min for color development and measure
          sample absorbance at  650 nm.

t    Calculations
     •Prepare a standard curve by  plotting  absorbance of the standards
          vs.  sulfide  concentration.   Ensure that  the  standards  cover the
          range  of  concentrations expected in  the  samples.   Determine the
          sulfide concentration of the sample distillate  by comparing sample
          absorbance with  the standard curve.
          Calculate total  sulfides concentration as follows:


              Concentration (mg/kg dry  weight) =


          Where:

                  C =  sulfide concentration in distillate,  mg/L
              0.05 =  sample volume of distillate,  L (as  written)
                  g =  wet  weight of sediment aliquot, g
                 %S =  percent solids of  sediment as a decimal fraction.

QA/QC Procedures

     It  is critical   that each sample  be thoroughly  homogenized  in the
laboratory before a  subsample  is  taken for  analysis.  Laboratory
homogenization should  be conducted even if samples were homogenized in the
field.
                                      35

-------
                                            Conventional Sediment Variables
                                                             Total Sulfides
                                                                 March 1986

     It is recommended that  triplicate analyses be conducted on one of every
20  samples,  or  on  one  sample  per batch  if  less  than  20  samples  are
analyzed.   Fresh standards should be used  to  calculate a calibration curve
for each batch of samples.   The  analytical  balance should be inspected daily
and calibrated at least once per week.

DATA REPORTING REQUIREMENTS

     Total sulfides should  be  reported  as mg/kg of  sediment  dry  weight to
the nearest  0.1 unit.   The laboratory  should  report  the  results  of  all
samples (including QA  replicates)  and should note any problems that may have
influenced  sample quality.   The  laboratory  should  also  describe  the
calibration curve used to determine  total  sulfide concentrations.
                                      36

-------
                                            Conventional Sediment Variables
                                                             Total Nitrogen
                                                                 March 1986
                              TOTAL NITROGEN
     If the elemental analyzer used  to measure total organic carbon can also
measure  total  nitrogen,  it  is  recommended that  the latter  variable  be
measured simultaneously  with TOC.   Total  nitrogen  values  in sediments
generally are used  to compare carbon-to-nitrogen  ratios.   A separate total
nitrogen analysis using a technique  other  than the elemental analyzer method
is not considered equivalent  for calculating  carbon-to-nitrogen ratios.
                                      37

-------
                                             Conventional  Sediment  Variables
                                                   Biochemical  Oxygen Demand
                                                                 March  1986
                      BIOCHEMICAL OXYGEN DEMAND (BOD)
USE AND LIMITATIONS
     Biochemical oxygen demand is a measure of the dissolved oxygen  consumed
by microbial organisms while  assimilating  and oxidizing the organic  matter
in a  sample.   This test  is  used to estimate  the amount of organic  matter
that is available to organisms, in contrast to other tests  used  to  estimate
the  total  amount  of organic  matter  (e.g., total  volatile solids,  total
organic carbon, chemical  oxygen demand).

     In addition  to oxygen used  for degrading organic  matter,  biochemical
oxygen demand  may also  include  oxygen used  to  oxidize inorganic  material
(e.g., sulfide, ferrous iron)  and reduced forms of nitrogen.

     The method  described by  Plumb  (1981) involves the use of  freshwater
bacteria as  the seed and  buffered  distilled  water as  the  dilution water.
An alternative procedure  for  marine  sediments is  to use marine  bacteria  as
the  seed  and  filtered,  oxygenated seawater as  the  dilution  water.   If
seawater is used, it does not need to be buffered.

FIELD PROCEDURES

Collection

     Samples can  be collected  in glass or  plastic containers.  A  minimum
sample size of  50 g  is  recommended.   If unrepresentative material  is  to  be
removed  from  the  sample,   it  should be  removed  in  the  field under  the
supervision of the chief scientist and noted on the field log sheet.

Processing

     Samples should  be stored  at 4°  C, and  can  be held  for  up to  7 days
under that  condition.   Samples should  be  kept field-moist and  air  contact
should be prevented to minimize oxidation.

LABORATORY PROCEDURES

Analytical Procedures

t    Equipment
     -Incubator
               Thermostatically controlled at 20° +_ \o  c.
               All  light  should  be  excluded  to  prevent the  photosynthetic
               production of  dissolved oxygen by algae  in the sample.


                                      38

-------
                                       Conventional Sediment Variables
                                             Biochemical Oxygen Demand
                                                            March 1986

     Incubation  bottles
          300-mL capacity, with ground glass stoppers.
     Phosphate buffer solution
          Dissolve  the  following  in  distilled  water:   8.5  g potassium
          dihydrogen  phosphate,  K^POy,  21.75  g  dipotassium hydrogen
          phosphate,  ^HPtty;  33.4 g disodium hydrogen phosphate hepta-
          hydrate,  N32HP04-7H20;   and  1.7 g ammonium  chloride,  NH/iCl.
          Dilute to 1 L.   The pH of this  buffer  should  be  7.2  without
          further adjustment.  If  dilution water  is to be stored in the
          incubator,  the  phosphate  buffer should be added just prior to
          using  the dilution  water.
     Magnesium sulfate solution
          Dissolve  22.5 g MgS04«7H20  in distilled water and dilute to 1
          L.
     Calcium chloride solution
          Dissolve  27.5 g anhydrous CaCl2  in distilled water and dilute
          to 1 L.
     Ferric chloride solution
          Dissolve  0.25 g FeCl3«6H20  in distilled water and dilute to 1
          L.
     Dilution water
          Store  distilled  water  in cotton-plugged bottles for  a
          sufficient length  of  time to become  saturated with dissolved
          oxygen.  The  water should  be aerated by shaking  a partially
          filled bottle or using  a  supply  of clean compressed air.  The
          distilled water used should be  as near as  possible to 20° C
          and of high purity.   Place the desired volume of distilled
          water   in  a  suitable  bottle and  add  1 ml  each  of phosphate
          buffer, magnesium  sulfate, calcium  chloride,  and  ferric
          chloride  for each  liter of  water.
     Seeding material
          Satisfactory   seed may sometimes be obtained by  using the
          supernatant liquor from domestic sewage that has  been stored
          at 20° C  for  24-36 h.   Use the  seed that  has been found by
          practical experience to be the  most  satisfactory for  the
          particular material under study. Only past  experience can
          determine the   amount of  seed to be  added per  liter  but the
          amount should  give an  oxygen  depletion of approximately  2
          mg/L.   The amount  of seed  required may vary with the source
          of the seed.    Seeded dilution  water  should  be used the same
          day it is prepared.

Standards preparation
-    Prepare a  stock  BOD   standard solution by dissolving  0.150  g
     reagent grade  glucose and 0.150  g reagent  grade glutamic acid in 1
     L  of distilled water.  The solids  should  be dried  for  1  h  at
     103° C prior to weighing.
                                 39

-------
                                            Conventional Sediment Variables
                                                  Biochemical  Oxygen Demand
                                                                 March 1986

•    Sample preparation
     -Allow samples  to warm  to room temperature.
          Homogenize  each  sample mechanically,  incorporating  any  overlying
          water.
          Remove a representative aliquot  (approximately 25 g)  and  analyze
          for total solids content.
          Remove a representative aliquot  (approximately 5 g) and weigh  it
          to the nearest 0.1 mg.
          Transfer the weighed aliquot to a BOD bottle for analysis.

•    Analytical procedures
     -Fill each BOD  bottle  with  dilution  water and place  the samples  in
          the incubator.   Ensure that air bubbles are not trapped in  the BOD
          bottles.   Prepare  a blank consisting of  dilution  water  in  a
          separate BOD bottle.   Make  sure that  there  is  a water  seal  in the
          neck  of  each  sample  bottle  and  blank  when  placed  in the
          incubator.   Replenish  the water seals on all bottles  each morning.
          Determine  the  initial  dissolved  oxygen  concentration of  each
          sample and  blank using the Winkler titration method or  a dissolved
          oxygen probe.   This can best be accomplished by directly measuring
          the dissolved  oxygen  concentration  in  the dilution water.   This
          method  is  recommended  because  sediment  may  cause a  rapid
          consumption  of  oxygen,  making  it  difficult  to obtain  a   stable
          initial dissolved oxygen reading.  If a probe  is used for oxygen
          measurement, the same sample can be  used for  immediate dissolved
          oxygen demand  and biochemical oxygen  demand.
          Incubate samples and  blanks for  5  days at 20  +_  1°  C.  Determine
          residual dissolved oxygen  concentrations  in  the incubated  samples
          using  the  analytical method  of  choice.   The most reliable  BOD
          determinations  will   occur in  those  samples with  a residual
          dissolved oxygen concentration  of at  least  2 mg/L  and  a dissolved
          oxygen depletion of at least 2 mg/L.
          It is recommended that the dilution water be incubated as  a check
          on its quality.  To do  this,  fill  two BOD  bottles  with  unseeded
          dilution water.  Stopper  one  bottle,  fill  the  water seal,  and
          place  in the  incubator at  20  +_ 1° C  for  5  days.   Analyze the
          second  sample  to   determine   initial  dissolved  oxygen
          concentration.  Following  the  5-day  period,  determine  dissolved
          oxygen in the  incubated sample.  The oxygen  depletion should not
          be more than 0.2 mg/L and  preferably not  more than  0.1 mg/L.  If
          these values  are  exceeded,  the quality of the dilution water  or
          the treatment  of samples  (e.g., filling of water  seals) should  be
          considered  suspect.
          Prepare a working BOD standard solution by  diluting 20 mL of the
          stock solution to 1  L with seeded dilution  water.   Fill three BOD
          bottles and incubate at 20 + 1°  C  for 5 days.  The resulting BOD
          of  these samples should Fe  218  *_  11  mg/L.  Any  appreciable
          deviation from these expected results may raise questions  on the
          quality of  the dilution water,  the  viability or suitability of the
          seed material, or the  analytical  technique.

                                      40

-------
                                            Conventional Sediment Variables
                                                  Biochemical Oxygen Demand
                                                                 March 1986
•    Calculations
     -Sediment BOD is calculated  as  follows:

               BOD (mg/kg dry weight)  =


          Where:

                0 = dissolved oxygen  concentration at time zero, mg/L
                F = dissolved oxygen  concentration after 5 days, mg/L
                b = volume of BOD  bottle,  ml
                g = wet weight  of  sediment sample used, g
               %S = percent solids in sediment  sample (expressed as a
                    decimal fraction.

QA/QC Procedures

     It  is critical  that each  sample  be  thoroughly homogenized  in  the
laboratory before  a  subsample  is taken  for  analysis.   Laboratory
homogenization should be  conducted even  if samples  were  homogenized in the
field.

     It is recommended that triplicate analyses be conducted on one of every
20  samples,  or  on  one sample per  batch  if  less than  20  samples  are
analyzed.  A dilution water blank and glucose-glutamic acid standard should
be analyzed at the same frequency  as  the  triplicate  analyses.

DATA REPORTING REQUIREMENTS

     Biochemical oxygen  demand  should be reported  as  mg/kg  dry  weight, to
the  nearest  0.1 unit.   The laboratory  should report  the  results  of  all
samples  analyzed,  including QA  replicates,  seeded dilution  water  blanks,
unseeded dilution  water blanks,  and  glucose-glutamic  acid  standards.   The
laboratory should  also note any  problems that may have  influenced sample
quality.
                                      41

-------
                                             Conventional Sediment Variables
                                                      Chemical  Oxygen Demand
                                                                  March 1986
                       CHEMICAL OXYGEN DEMAND  (COD)
USE AND LIMITATIONS
     Chemical  oxygen  demand is  a measure of  the oxygen equivalent  of  the
organic matter  content  of a sample  that is susceptible  to oxidation by  a
strong chemical  oxidant at elevated  temperature  and reduced pH.   The  test
was devised  as an  alternative  to the  biochemical  oxygen  demand  test  for
estimating organic  matter.  For samples from  a  specific  source,  chemical
oxygen demand can be related empirically to biochemical  oxygen demand, total
organic carbon, or  total  volatile solids  and  then used  for monitoring after
a relationship has been established.

     Major limitations of the chemical oxygen demand test are that it is  not
specific  for  organic matter and that correlations  with other  measures  of
organic carbon  are not  always  found.   Inorganic substances such  as Fe^+,
MnZ+t  and S2" can  increase the  consumption  of  oxidizing  agent during  the
test.  Plumb  (1981) recommends  that  chemical  oxygen demand not be equated
with organic matter in sediments.

FIELD PROCEDURES

Collection

     Samples  can  be collected  in glass  or  plastic containers.   A minimum
sample size of 50 g is  recommended.   If unrepresentative material  is to be
removed   from the  sample,  it  should  be  removed  in  the  field  under  the
supervision of the chief scientist and noted on the field log sheet.

Processing

     Samples  should be  stored  at 4°  C  and  can  be held  under that condition
for  7  days.   Samples  must be  kept  field-moist  and  free from  air contact
during storage to prevent air oxidation.
                                      42

-------
                                            Conventional Sediment Variables
                                                     Chemical Oxygen Demand
                                                                 March 1986

LABORATORY PROCEDURES

Analytical Procedures

•    Equipment
     •Reflux apparatus
               Consisting of 250-  or 500-mL Erlenmeyer  flasks  with ground
               glass  24/40  neck1  and  300-mm  jacket  Liebig,  West,  or
               equivalent condensers2 with 24/40 ground glass joint.
          Hot plate
               Having  sufficient  power  to  produce  1.4  W/cmz (9  W/inz)  of
               heating surface,  or equivalent,  to  ensure adequate refluxing
               of the  sample.
          Standard potassium dichromate  solution, 0.250 H_
               Dissolve 12.259  g KgCrgOy primary standard grade, previously
               dried at 103° C  for  2  h, in  distilled water and  dilute to 1
               L.  The addition  of  0.12 g/L sulphamic acid  will  eliminate
               interference due to nitrites  in the sample at concentrations
               up to 6 mg/L.
          Sulfuric acid reagent
               Concentrated H2S04 containing 22  g silver  sulfate,  Ag2S04,
               per 9-lb bottle.   Allow  1 or  2 days for dissolution.
          Standard ferrous  ammonium sulfate  titrant, 0.25 N_
               Dissolve  98  g   Fe(NH4)2  (SO/^'SHgO  in  distilled water.
               Carefully  add 20 ml concentrated HgSO^ cool, and  dilute to 1
               L.   This  solution  must be  standardized  against  I^C^Oy
               daily.   To standardize the ferrous  ammonium sulfate, dilute
               10 ml standard potassium dichromate solution to approximately
               100 mL.  Carefully add 30 ml  concentrated H2S04 and allow to
               cool.  Titrate with ferrous ammonium titrant, using 2-3 drops
               of ferroin indicator.

                                             (ml K?Cr?07)(0.25)
                                 Normality = 	1—LJ.	
                                             [ml Fe(NH4)2(S04)2]

          Ferroin indicator
               Dissolve  1.485  g 1,10-phenantroline  monohydrate  and 0.695 g
               ferrous sulfate,  FeS04«7H20,  in water and dilute to 100 ml.
               Alternatively, a  commercially  prepared  indicator can  be
               purchased.
          Silver sulfate
               Ag2s°4i reagent  powder.
          Mercuric sulfate
               HgS04,  analytical-grade crystals.
     iCorning 5000 or equivalent.

     2Corm'ng 2360, 91548, or equivalent.

                                      43

-------
                                            Conventional Sediment Variables
                                                     Chemical  Oxygen  Demand
                                                                 March  1986

•    Sample preparation
     -Allow samples  to warm  to room temperature.
          Homogenize  each sample mechanically,  Incorporating  any  overlying
          water.
          Remove a representative aliquot  (approximately 25 g)  and analyze
          for total solids content.
          Remove a representative aliquot  (approximately 2 g) and weigh  it
          to the nearest 0.1 mg.
          Transfer the weighed aliquot to  a reflux  flask for  COD analysis.
          Wash the sediment  into  the  flask  with  a  minimum  amount  of
          distilled water (i.e.,  <25 ml).

t    Analytical procedures
     -Place several  boiling  stones or glass beads and 1.0 g HgS04  in the
          reflux flask with the  sample.
          Add 25 ml 0.25 ^ <2CR207 to the flask and mix thoroughly.
          Slowly, and  with  constant  mixing,  add  75  ml  of  sulfuric
          acid-silver  sulfate solution.  Ensure that  the  mixture is  well
          mixed to avoid localized superheating.
          Attach the sample flask to a condenser and reflux  for 2 h.   If the
          added dichromate dissipates during reflux, either 1)  repeat, using
          a  smaller  sample   size,  or 2)  carefully  add additional  0.25 N_
          K2C"r2°7  through  the  condenser.   Be sure  to  record any  added
          dichromate.
          Allow the sample to cool and  rinse  the condenser  with 40-50  ml
          distilled water.
          Add an additional 50 ml of distilled water to the sample and allow
          to cool  to room temperature.
          Add  3-5 drops of  ferroin  indicator  and  titrate with  0.25  N^
          Fe(NH4)2(504)2  to  a  sharp   color  change   (blue-green  to
          reddish-brown).
          For  a  blank,  reflux 25 ml  of distilled water,  25  ml of  0.25 N_
          K2Cr2°7» *  Q HgS04,  several  glass beads  or boiling stones,  and 75
          ml of sulfuric acid-silver  sulfate  solution  for  2  h.  Cool,  add
          3-5 drops of ferrion indicator, treat as a sample and titrate with
          0.25 N_Fe(NH4)2(S04)2.

•    Calculations
     -Sediment COO is determined as follows:

                         COD (mg/kg dry weight) = 'A'

          Where:
               A = volume of 0.25 N Fe(NH4)2(S04)2 for blank titration,  mL
               B = volume of 0.25 FT Fe(NH4J2(S04)2 for sample titration, ml
               N = normality  of (TT25  N^ Fe(^4)2(504)2  used  for titration,
                   eq/L
           8,000 = equivalent  weight of oxygen, mg/eq
                                     44

-------
                                            Conventional Sediment Variables
                                                     Chemical  Oxygen  Demand
                                                                 March  1986

               g =  wet weight of sample, g
              %S =  percent  solids  in sediment sample (expressed as a decimal
                   fraction.

QA/QC Procedures

     It  is  critical  that each  sample be  thoroughly  homogenized  in  the
laboratory  before a  subsample   is  taken  for  analysis.   Laboratory
homogenization should  be conducted even if samples were  homogenized  in  the
field.

     It is recommended  that triplicate analyses be conducted on one of every
20  samples,  or on  one  sample  per  batch  if  less  than  20   samples  are
analyzed.

DATA REPORTING REQUIREMENTS

     Chemical   oxygen demand should be  reported  as mg/kg  of  sediment  dry
weight to the  nearest 0.1 unit.  The laboratory should report the results of
all samples analyzed (including QA replicates and method blanks)  and  should
note  any  problems  that may have  influenced  data quality.  The  laboratory
should also report  the  results  of  the standard test.
                                     45

-------
                                             Conventional  Sediment  Variables
                                                                  References
                                                                  March  1986
                                REFERENCES

Buchanan, J.B.  1984.   Sediment analysis,  pp.  41-65.   In:   Methods  for
the Study of Marine Benthos.   N.A.  Holme and  A.D.  Mclntyre (eds).   Blackwell
Scientific Publications, Boston, MA.
Folk,  R.L.   1968.   Petrology  of  sedimentary rocks.   Hemphill  Publishing
Co., Austin, TX.  172 pp.
Hedges,  J.I.,  and  J.H. Stern.   1984.  Carbon and nitrogen determinations
of carbonate-containing solids.  Limnol. Oceanogr.  29:657-663.
Krumbein, W.C. and F.J. PettiJohn.   1938.  Manual  of sedimentary petrography.
Appleton-Century Crafts, NY.   549 pp.
Plumb, R.H.   1981.  Procedures for  handling and chemical analysis of sediment
and water samples.  Technical Report EPA/CE-81-1.  U.S. Army Corps of Engineers,
Vicksburg, MS.
U.S.  Environmental Protection Agency.  1983.  Methods for chemical analysis
of water and  wastes.  EPA 600/4-79-020  (Revised 1983).  Environmental Monitoring
and Support Laboratory, Cincinnati, OH.
                                     46

-------
ORGANIC COMPOUNDS

-------
METALS

-------
FINAL REPORT                        Pi/qef Sound Estuary Program
TC-3090-04
RECOMMENDED PROTOCOLS
FOR MEASURING METALS
IN PUGET SOUND WATER, SEDIMENT,
AND TISSUE SAMPLES
Prepared by:
TETRA TECH, INC.

Prepared for:
RESOURCE PLANNING ASSOCIATES

for:
PUGET SOUND DREDGED DISPOSAL
ANALYSIS (PSDDA)
Monitored by:
U.S. Army Corps of Engineers
Seattle District
Seattle, WA
August, 1986
TETRA TECH, INC.
11820 Northup Way
Bellevue, WA 98005

-------
                                 CONTENTS


                                                                        Page

LIST OF TABLES                                                           iv

ACKNOWLEDGMENTS                                                          iv

INTRODUCTION                                                              1

    GENERAL                                                               1
    PRECAUTIONS AND LIMITATIONS                                           1

SAMPLE COLLECTION                                                         4

    PRECOLLECTION                                                         4
    SOURCES OF CONTAMINATION                                              4
    WATER COLUMN SAMPLES                                                  6
    PARTICULATE SAMPLES                                                   7
    SURFICIAL SEDIMENT SAMPLES                                           10
    TISSUE SAMPLES                                                       13

ANALYTICAL METHODS                                                       16

    SAMPLE PREPARATION                                                   16
    INSTRUMENTAL METHODS                                                 18
    RECOMMENDED INSTRUMENTAL METHODS                                     18
    METALS SPEC IAT ION                                                    19

QUALITY ASSURANCE/QUALITY CONTROL                                        20

    QA/QC MEASURES  INITIATED BY THE ANALYTICAL LABORATORY                20
    QA/QC MEASURES  INITIATED IN THE FIELD                                21
    CORRECTIVE ACTIONS                                                   23

DATA REPORTING                                                           24

    DATA REPORT PACKAGE                                                  24
    BACKUP DOCUMENTATION                                                 24

REFERENCES                                                               25

   APPENDICES

       A:  SELECTION OF METALS FOR PROTOCOL DEVELOPMENT                 A-l

       B:  EXCERPTS FROM EXHIBITS B, D, E, AND G (U.S. EPA 1985)        B-l

       C:  ELUTRIATE AND FRACTIONATION METHODS (PLUMB 1981)             C-l
                                     11

-------
D:  TOTAL ACID DIGESTION METHOD FOR SEDIMENT (RANTALA AND
    LORING 1975)                                                 D-l

E:  HN03/HC104 DIGESTION METHOD FOR TISSUE (TETRA TECH 1986a)    E-l

F:  APDC/MIBK EXTRACTION METHOD FOR SALT WATER (GREENBERG ET AL.
    1985)                                                        F-l

G:  DFAA INSTRUMENTAL AND SPECTROPHOTOMETRIC METHODS (U.S. EPA
    1979a)                                                       G-l
                              111

-------
                                 TABLES


Number                                                               Page

  1     Contributors to the  metals protocols                             2

  2     Sample preservation  and storage parameters                       5
                             ACKNOWLEDGMENTS


     This chapter was  prepared by Tetra Tech,  Inc.,  under the direction
of Mr.  Robert  Barrick, for Resource Planning Associates (RPA) for the Puget
Sound  Dredged Disposal Analysis  (PSDDA) monitored  by the Seattle District,
U.S. Army  Corps of Engineers  (COE)  in partial fulfillment  of Contract
No. DACW67-85-D-0029.  Dr.  David Kendall  of U.S. COE was the Technical
Coordinator.  The primary  author of this chapter was  Dr. Charles Lytle
of Tetra Tech,  Inc.
                                   1v

-------
                                                                    Metals
                                                              Introduction
                                                               August  1986
                               INTRODUCTION
GENERAL

     This chapter  presents recommended protocols for measuring metals  in
water, sediment,  and tissue samples from Puget Sound.  The  list of metals
considered and  their limits of detection (LODs) are summarized in Appendix  A.
The main objectives of this report are to:

     •    Elucidate acceptable method  options for each metal  meeting
          the minimum LODs specified for water, sediment, and biological
          samples from Puget Sound

     •    Prioritize all method options for each metal and make recom-
          mendations on the best method for each appropriate matrix.

     The  recommendations presented in this chapter are based on the  results
of a workshop and  detailed reviews by representatives from most organizations
that  fund or conduct environmental  research  in Puget  Sound (Table 1).
The purpose of  developing these recommended  protocols  is  to encourage all
Puget  Sound investigators conducting monitoring programs, baseline surveys,
and intensive investigations to use standardized methods whenever possible.
If  this  goal  is achieved,  most data  collected  in Puget  Sound should  be
directly comparable  and thereby capable of being integrated into a sound-wide
database.  Such a database  is necessary for developing  and maintaining
a comprehensive water quality management program for Puget Sound.

     Although  the  following protocols are recommended for most  studies
conducted in Puget  Sound, departures  from  these  methods  may be necessary
to meet the special requirements of individual projects.  If such departures
are made, however,  the funding agency or investigator should be aware that
the resulting  data may not be comparable with most other data of that kind.
In some instances,  data collected using different methods  may be compared
if the methods  are  intercalibrated adequately.

PRECAUTIONS AND LIMITATIONS

     Although  well-established  procedures for proper sample handling and
analysis are essential for collecting data of high quality, standard protocols
do  not exist for all the  required methods.  Approved methods exist for
the analysis of  metals  in water  (e.g., U.S.  EPA 1979a,  ASTM 1983), but
only  interim procedures have been proposed  for sediment and tissue samples
(U.S.  EPA 1977, Plumb 1981).  New considerations  may need  to be addressed
and additional  guidelines  provided to maximize the quality and efficiency
of the procedures.

-------
    TABLE  1.  CONTRIBUTORS TO THE METALS PROTOCOLS
Roy Arakia
Bob Barricka«b
Robert Clark3
Eric Crecelius3
Ray Dalsega
Bob Dexter3
Mark Fugiel3
Michael Higgins3
Roger Kadeg3
David Kendall3
Mary Beth Lanza3
Chuck Lytle3
Pamela Navaja3
Matt Schultz3
Mike Shelton
Jeff Stern3
Steve Twiss3
Washington Dept. of Ecology
Tetra Tech
NOAA/NMFS
Battelle Northwest
Metro
EVS Consultants
Am Test
Analytical Technologies
Envirosphere
U.S. COE
Laucks Testing Labs
Tetra Tech
Washington DSHS
Envirosphere
Weyerhaeuser Co.
Tetra Tech
Washington Dept. of Ecology
3 Attended the workshop held on January 17, 1986.
b Workshop moderator.

-------
                                                                    Metals
                                                              Introduction
                                                               August  1986
     The instrumental  methods proposed are prone  to  chemical  and matrix
interferences  that can either suppress  or enhance the analyte signal  (Skoog
1985).   Special attention should be given when they are likely to  be present
as noted in the individual  protocols.   If  any such interferences are  suspected,
it is extremely important  that  their  effect be determined and  corrective
action  taken.

     Because  of  the low limits of detection  proposed for many environmental
analyses, clean field and laboratory procedures  are especially important.
Sample  contamination can occur  during  any stage of collection, handling,
storage, or  analysis.   Potential contaminant  sources must be  known and
steps taken  to minimize or eliminate them (Murphy 1976).

-------
                                                                   Metals
                                                        Sample Collection
                                                              August 1986
                           SAMPLE COLLECTION
PRECOLLECTION

Container Materials

     The best containers for the collection and/or storage of trace metal
samples are  made of quartz  or tetrafluoroethylene  (TFE).  Because  these
containers  are  expensive,  the preferred container material is either poly-
propylene or linear polyethylene, with a polyethylene cap.  Borosilicate
glass  may be  used, but  lids  must  not have aluminum or cardboard liners
(Greenberg et  al. 1985).

     For tissue samples, scalpels  should  be  made  of corrosion-resistant
stainless steel, while  tweezers and cutting  surfaces should be  plastic
or TFE  (Tetra Tech 1986a).   Stainless steel  homogenizer blades should be
replaced with  tantalum.

     The recommended container materials, sample sizes, preservation techniques,
and storage  lifetimes  for all  the metals of concern in water, sediment,
and tissue are summarized in Table 2.

Cleaning Methods

     Prior  to use,  containers  and  any  glass  or plastic parts associated
with the sampling equipment should be thoroughly  cleaned with a detergent
solution, rinsed with  tap  water,  soaked 24 h at 70° C in an acid  solution
of 1:1  HN03 or  1:1 HC1, and  then  rinsed with  metal-free water.  Do not
use chromic  acid for  any cleaning purpose.   For stainless steel  parts,
omit the acid  soaking  step (Greenberg et  al. 1985).

SOURCES OF CONTAMINATION

     In the  field,  sources of contamination include sampling gear, lubricants
and oils, engine exhaust, airborne dust,  and ice used for  cooling  samples.
During sample handling and analysis, contamination sources include  exposure
to airborne  dust,  insufficiently clean sample  containers, contact with
inappropriate materials (e.g., rubber  o-rings, aluminum-lined caps, steel
containers), contaminated reagents,  and  carryover in testing instruments
due to insufficient cleaning or flushing  between samples.

     Exposure to  airborne dust can be minimized by  using capped containers
and by keeping physical  sample handling to a  minimum.  Use of recommended
container materials  and cleaning  methods should  minimize contamination
from storage containers.  Reagents should be ACS Reagent  Grade or better
and should never  be returned to their stock containers once removed.  Carry-
over can be avoided  by  carefully following instrument  manufacturers'  instruc-

-------
                   TABLE 2.  SAMPLE PRESERVATION AND STORAGE PARAMETERS
Sample
Analyte
Total or dissolved
metals (except Hg)a
Total or
dissolved Hga
Particulate Metals
All metals
Elutriate Studye
Fractionation Studye
All metals (except Hg)
Hg
Matrix
H20
H20
H20
Sediment
H20
Sediment
Sediment
Tissue
Tissue
Containerb
P.G.TFE9
G.TFE
P.G
P,G
P|G
P,G
P,G
P,G
Size
1 L
250 ml
1 gal
50 gc
12 L
3 L
1 L
5 gc.h
0.2 gc,h
Storage
Preservative
HN03 PH <2
HN03 pH <2
i
Freeze
4° C
4° C
Freezef
Freezef

Lifetime
6 mo
28 days
i
6 mod
--
—
6 mod
28 daysd
a U.S. EPA 1982.
b P = linear polyethylene, G = borosilicate glass, TFE = tetrafluoroethylene.
c Wet weight.
d Suggested.  No EPA criteria exist.  Hg holding time 28 days.
e Plumb 1981.   Storage time "as short as possible," analyses to be completed
"within 1 wk of sample collection."
f Post-dissection.
g If aliquot for Hg taken  from this  1 L sample, cannot use  linear polyethylene.
h Weight  is a minimum for  one sample.  Studies using specific  organs may  require more.
i Samples should be filtered as soon as possible and always within 24  h.  Workshop attendees
recommend that  filtering be done  shipboard  rather  than  in  lab  on  shore.

-------
                                                                   Metals
                                                         Sample Collection
                                                              August 1986
tions.   Sources of contamination in the analytical laboratory and on shipboard
have been thoroughly reviewed by Mart  (1979a,b).

WATER COLUMN SAMPLES

     Water  column  samples frequently are collected using water bottles.
Water bottle samplers are relatively simple devices that generally  consist
of some  type of  cylindrical  tube with stoppers at each end  and a closing
device that  is activated by a messenger or an electrical signal.  The most
commonly used  samplers of  this description  are  the Kemmerer, Van Dorn,
Niskin, and  Nansen samplers.   Each device  samples a discrete  parcel  of
water  at any designated depth.  Frequently, multiple water samplers are
fixed on a rosette frame so that several depths can be sampled during one
cast or replicate samples can be taken at the same depth.

     Prior  to deployment, the stoppers of water bottle samplers  are cocked
open on the  sampling vessel.  At this  step, it  is critical  that the interior
of the  sampler  and stoppers  remain  free from contamination.  All members
of the sampling team should therefore avoid  touching the insides  of the
sampler and  stoppers, and all samplers and  stoppers must be rinsed  thoroughly.

     After  cocking,  the sampler  is  lowered to a designated depth.  The
sampler must be open at both ends so that water is  not trapped within the
device  as it is  being  lowered through the water column.  Once the sampler
reaches the  desired depth, it should be allowed to equilibrate  with  ambient
conditions for 2-3 min before being closed.

     After  equilibration, the closing device  can be activated by messenger
or electrical  signal, and the sampler  can be  retrieved.  It is  recommended
that at  least two samplers be used simultaneously for each depth.  A second
sampler will provide a  backup to the primary  sampler in case the  latter
device  misfires  or won't  trigger.  This will  eliminate the need  for an
additional cast.  A second sampler will also  supplement the primary  sampler
if the  volume collected by the latter device  is too small for all required
subsampling  and rinsing.   To ensure that  all  subsamples at  a particular
depth  are collected from  the same water  parcel, it is essential that they
all  be taken  from a single cast.  Multiple casts using a single  water sampler
will not meet this objective.  Sample  water must therefore be used conserva-
tively after collection.

     Once the water  sampler  is brought  on board the sampling vessel, the
stoppers should be checked immediately for complete  seals.   If  a  stopper
is not  properly sealed,  water from the  sampled depth may  have  leaked out
upon retrieval and been  replaced by water  from shallower  depths.  Because
this  kind of contamination can bias results, the entire water  sample  should
be rejected.

-------
                                                                   Metals
                                                         Sample Collection
                                                              August 1986
PARTICULATE  SAMPLES
     This section  provides guidance on  the  collection of the particulate
fraction  of water samples  from  in-plant process  streams, effluents,  and
receiving waters.   All devices  capable of separating  particulate material
from water  samples may contribute contaminants  to the sample taken  for
analysis.  The  level  of this  "blank"  contamination must be determined in
advance and reduced when possible.

     There are  several  techniques  for obtaining the  particulate fraction
of water samples,  including different  filtration procedures, settling/centrifu-
gation,  centrifugation, and continuous-flow centrifugation.  The latter
technique may  not be appropriate for  metals because the  equipment  required
contains a number of metallic parts.   A backflush-filtration technique
has  also  been  described in an intercomparison  study (Horowitz 1986),  but
is not appropriate for chemical  analyses because of selective losses of
the  filtered material prior to analysis.

     Filtration may  be prefereable  to other techniques  for many analyses
because it  requires less expensive equipment (i.e., a high  speed centrifuge
is not required), and provides a sample that is suitable for direct chemical
analysis  (i.e.,  a residue on a filter that can  be  extracted or digested).
The amount of particulate material that may be collected on a filter (e.g.,
typically <1 g)  may limit this technique for low-level analyses, including
metals.  Centrifugation techniques can yield comparable results and may
be used to  collect much  larger amounts of particulate material  than by
filtration (e.g.,  several grams),  but require careful and  complete transfer
of the sample  from  the centrifuge prior to analysis.

     Horowitz (1986)  describes a centrifugation  technique (fixed angle-
head, or swing-bucket rotor; not a continuous-flow system) for metals analysis
that  he  recommends when  large-scale studies are undertaken (e.g., 2,000-
3,000 samplers per  year).   Bates et  al.  (1983)  describe a system  using
continuous-flow  centrifugation (Son/all Model SS-3 or RC-5  high-speed centri-
fuge)  for collection of particulate material  for organic analyses (specifically
hydrocarbons).   One interpretation of the results of this  study that may
have implications  for metals analyses was  the possibility that continuous-
flow  centrifugation may  not  recover fine-grained particles in the l-2um
range as  efficiently as  filtration.  Although both techniques recovered
between  70 and  90 percent  of  the particles in this size  range, these very
small particles  may have a higher loading  of metals than larger particles.

Sample Compositing  and Preservation

     Ideally, samples of in-plant  process streams or effluents  should be
collected as a flow-proportionate composite of at least 5 days of mechanical
interval  sampling  to yield a time-integrated measurement.   If flow-proportion-
ated samples cannot be collected (i.e., aliquots  collected  in  proportion
to the changing flow  rate  of  the stream), instantaneous grab  samples can

-------
                                                                   Metals
                                                         Sample Collection
                                                              August 1986
be substituted, but should still  be composited over several days collection
whenever possible.  It is usually  unreasonable to composite  receiving  water
samples  over  such periods of  time  in a routine  sampling program, and it
may be unreasonable to store composited process stream or effluent  samples
over  several  days when total  volumes greater than approximately 20 L (i.e.,
5 gal) are required.

     Samples  for  metals analysis should  be stored under  refrigeration in
1-gal  polyethylene cubitainers.  Until  use, the cubitainers  should be stored
(collapsed)  in plastic bags.  The  cubitainers should  be  carefully rinsed
with several aliquots of the  sample prior to  filling to remove any dust
particles.  An acid pre-wash  step is not  considered  necessary  for the
cubitainers.

     Filtration of  the  entire sample should be conducted  at one time, and
this process may take from one to  several hours  to complete.  A potential
preservation technique is addition of mercuric chloride (for organic analyses
only), or a 2  percent sodium azide solution.  The latter preservative has
worked  well with sediment trap  samples collected by  NOAA, but requires
access to a fume hood for sample processing which is a distinct limitation.
Suspended solids  samples cannot be preserved with 100 percent effectiveness
(Greenberg et  al.  1985).  Hence, preserved sample bottles should be  stored
under  refrigeration (2°C) prior to filtration to minimize decomposition
of solids. No samples should be held for longer  than 7 days.

Sample Volumes

     A 4-L composited  sample (i.e., 1-gal)  is recommended  for the analysis
of metals in  particulate  samples.  The volume  required for analysis is
a  function of the concentration of particulate material  in  the stream and
the concentration  of metals in the particulate  material.  Horowitz  (1986)
recommends the collection of  100 mg  of particulate material for analysis
of metals, but this quantity cannot always be obtained with a 4-L  sample.
Larger  sample volumes  (e.g., 20-L)  can be processed but are  not considered
necessary for  most purposes.   Smaller sample  sizes (e.g., 5-30 mg) have
been  successfully analyzed  for  most  metals using graphite furnace and ICP
techniques and a final dilution volume  of 10  mL digestate (Tetra Tech  1985).

Sample Processing

Preparation of Filters-

     Polycarbonate  membrane filters  (e.g., Nuclepore)  of  0.4-um pore  size
are recommended for filtration of  water samples for metals analyses.  The
142-mm  diameter filters  are required to filter the recommended minimum
of approximately 4-L of  sample.   The membrane  filters do not require any
pretreatment  prior to  weighing, and  can be taken directly  from the manu-
facturer's container.  Each filter must be carefully weighed to the  nearest
0.1 mg prior  to filtering.  Unlike glass  fiber filters, the  amount of residue


                                    8

-------
                                                                   Metals
                                                        Sample Collection
                                                              August 1986
collected  on  membrane filters  can be accurately weighed.  The dry weight
of the filtered material is used  directly to determine the concentration
of metals  associated with the particulate material.

     Because the 142-mm filters  are so light, care must be  taken to prevent
wrinkling  when placing the filter in the filter  holder.  Air currents  can
easily disrupt  the placement of the  filter, and once  wet  the filters must
be carefully moved to their final  position on the holder to  prevent tearing.
A TFE filter  holder is recommended,  but a plastic filter holder may also
be used.

Filtration of  Samples

     After the tared filter has  been placed into the filter  holder, a plastic
squeeze  bottle should be used to  completely rinse the filter with 10 percent
high  purity  HC1  followed by deionized,  distilled  water  until neutral pH
has been  established.   A water  aspirator can be used  to provide  suction
for the  rinsing.   The acid rinse will  leach any dust particles that may
have adhered to the filters (such particles can contribute  zinc and potentially
other contaminants to the analysis).

     Once  the filter  has been properly  rinsed, the  filter holder should
be reassembled and a siphon  tube attached between the filter holder  and
the cubitainer  holding the water sample (place the  cubitainer above the
filter holder  to provide gravity-feed).  The cubitainer  should be vigorously
shaken prior  to the filtration  to dislodge particles that  may have adhered
to the sides.  Using the water aspirator, siphon the entire water  sample
through  the  filter holder.   After the water has passed through the filter,
disconnect the filter  holder  from the aspirator while the aspirator is
still  running.

     Carefully open the filter  holder and, using  plastic forceps, fold
the filter in thirds  (like  a  crepe)  to protect the  filtered material.
Finally,  with two pairs  of  forceps gently hold down one end of the filter
while folding  the filter in half from the  opposite end.  Place the  folded
filter  in a  polycarbonate  petri dish.  Seal  the  petri dishes, label,and
store frozen  until the  metals analysis  can be started.  Prior to digestion,
the filters  should be dried  at no more  than 60°C and  weighed to determine
total  sample weight (the original  weight of the clean filter  must be supplied
to the laboratory).   Filters dried at  higher temperatures may suffer losses
of the more volatile metals.

     At least one  filter blank  should  be  prepared by going  through the
entire filtration setup procedure  (i.e., including the acid-rinse step)
in  the  absence of a water  sample.  Fold and  place the filter blank  in a
polycarbonate  petri  dish as with other  samples.   At least one filter blank
should  consist of an  untreated folded membrane  filter taken directly  from
the manufacturer's box  (to check efficiency of the rinse step).

-------
                                                                   Metals
                                                        Sample Collection
                                                              August  1986
SURFICIAL  SEDIMENT SAMPLES
     This section describes the  protocols required  to collect an acceptable
subtidal  surficial sediment sample  for subsequent measurement of physical
and chemical  variables.  This  subject has generally been neglected in the
past and  sampling crews have been given relatively wide latitude in deciding
how to collect  samples.  However,  because sample collection procedures
influence the  results of all  subsequent laboratory and  data analyses,  it
is critical  that samples be collected  using acceptable and standardized
techniques.

Design of Sampler

     In Puget Sound, the most common sampling device for subtidal surficial
sediments is the modified van  Veen bottom  grab.   However, a variety  of
coring devices  is also used.  The  primary criterion for an adequate sampler
is that it consistently collect  undisturbed samples to  the required depth
below the sediment surface without  contaminating the samples.  An additional
criterion is that the sampler can be  handled  properly  on board the survey
vessel.   An otherwise acceptable sampler may yield inadequate sediment
samples if it  is too large, heavy,  or awkward to be  handled properly.

     Collection  of undisturbed sediment requires that the sampler:

     •    Create a minimal bow wake when descending
     •    Form a leakproof seal  when  the sediment sample  is taken
     a    Prevent  winnowing and  excessive sample disturbance when
          ascending
     •    Allow  easy access to the  sample surface.

Most  modified van Veen  grabs  have open upper  faces that  are  fitted with
rubber flaps.   Upon descent, the  flaps are forced open to minimize the
bow wake, whereas upon ascent, the  flaps are  forced  closed  to prevent sample
winnowing.  Some box corers have solid flaps  that are  clipped  open  upon
descent  and snap shut after the corer  is triggered.  Although most samplers
seal adequately  when purchased, the wear and  tear of repeated  field use
eventually reduces this sealing  ability.  A  sampler  should  therefore be
monitored constantly for  sample leakage.  If  unacceptable leakage  occurs,
the  sampler should be repaired or  replaced.  If a sampler  is to  be borrowed
or leased for a  project, its sealing ability should be confirmed  prior
to  sampling.   Also, it  is  prudent to have a  backup  sampler  on board the
survey vessel  in case the primary sampler begins leaking  during  a  cruise.

     The required penetration depth below the sediment surface is  a function
of the desired  sample  depth (see Penetration Depth).   Generally,  it is
better to penetrate  below the desired sample depth to minimize  sample
disturbance when the  sampling device  closes.   Penetration  depth  of  most
sampling devices varies with sediment character, and generally is greatest
in fine sediments and least  in coarse sediments.  Sampling  devices generally
rely  upon either gravity or a piston mechanism to penetrate the sediment.


                                    10

-------
                                                                   Metals
                                                         Sample Collection
                                                              August 1986

In both  cases,  penetration  depth  can be modified by adding or subtracting
weight from  the  samplers.  Thus,  it  is optimal  to use a sampler  that has
a means  of  weight adjustment.   If a sampler  cannot consistently achieve
the desired  penetration depth, an alternate device should be used.

     Once the  sampler is secured  on  board the survey vessel, it is  essential
that the  surface of the sample be made accessible without disturbing the
sample.   Generally,  samplers have hinged flaps  on their upper face for
this purpose.  The opening(s) in  the upper face of the sampler  should be
large enough  to allow easy  subsampling of the  sediment surface.   If an
opening  is too small, the sample  may be disturbed as the scientific crew
member struggles to take a subsample.

Penetration  Depth

     For  characterizing surficial sediments in Puget Sound,  it  is  recommended
that the  upper 2 cm of the sediment  column  be  evaluated.  When collecting
the upper 2 cm  of sediment,  it is recommended that a minimum penetration
depth of  4-5 cm  be achieved for each acceptable  sample.

     Although  the 2-cm specification is arbitrary, it will  ensure  that:

     •    Relatively recent sediments are sampled
     •    Adequate volumes  of sediments can  be  obtained readily for
          laboratory analyses
     •    Data from different studies can be compared validly.

Sampling depths other than  2 cm may be appropriate for specific  purposes.
For example, the upper 1 cm of sediment may be required to determine the
age of the  most recently deposited sediments.  By contrast, a sample depth
much greater than 2 cm may be required to evaluate the vertical profile
of sediment characteristics or to  determine depth-averaged characteristics
prior to  dredging.  If a sampling depth other than 2 cm is used, comparisons
with data from 2-cm deep samples may be questionable.

Operation of Sampler

     The  sampling  device should be attached to the hydrowire  using a  ball-
bearing swivel.  The swivel will  minimize the twisting forces on the sampler
during deployment and ensure that proper contact is made with the bottom.
For safety,  the hydrowire,  swivel, and all  shackles  should have  a load
capacity at  least three times greater than  the weight of a full sampler.

     The  sampler  should be lowered  through the  water column at a  controlled
speed of  approximately 1 ft/sec.   Under no  circumstances should the sampler
be allowed  to  "free fall"  to the  bottom, as  this may result in premature
triggering,  an excessive  bow wake,  or improper orientation  upon contact
with  the bottom.   The sampler  should contact the bottom gently and only
its weight or  piston mechanism should be  used to force it into  the sediment.
                                    11

-------
                                                                    Metals
                                                         Sample Collection
                                                               August  1986

     After the  sediment sample is taken,  the  sampler should be raised slowly
off the bottom  and then retrieved at a controlled speed  of  approximately
1 ft/sec.   Before the sampler  breaks the water surface, the survey vessel
should head into  the waves (if present) to  minimize vessel  rolling.  This
maneuver  will  minimize  swinging of the sampler after it breaks the water
surface.   If excessive swinging occurs or if  the sampler strikes the vessel
during retrieval, extra attention should  be paid to  evaluating sample
disturbance when  judging  sample acceptability.

     The sampler should  be secured immediately after it is brought on board
the survey vessel.   If the  sampler tips or  slides around before being secured,
extra attention should be paid to evaluating  sample disturbance.

Sample Acceptability Criteria

     After  the sampler is secured  on  deck,  the sediment sample should  be
inspected carefully before being accepted.   The following  acceptability
criteria should be satisfied:

     •    The sampler is  not over-filled  with sample so that the sediment
          surface is pressed against the  top  of the sampler
     0    Overlying water is present (indicates minimal leakage)
     •    The overlying  water is not excessively turbid  (indicates
          minimal  sample  disturbance)
     •    The  sediment  surface is  relatively flat (indicates minimal
          disturbance or  winnowing)
     •    The  desired penetration depth is  achieved (i.e.,  4-5 cm
          for a 2-cm deep surficial sample).

If a sample does  not meet all criteria,  it  should be rejected.

Sample Collection

     After  a sample is judged acceptable,  the following observations should
be noted on the field log sheet:

     0    Station location
     0    Depth
     0    Gross characteristics of the  surficial sediment
               Texture
               Color
               Biological structures (e.g.,  shells, tubes, macrophytes)
               Presence of debris (e.g.,  wood  chips, wood  fibers,
               human artifacts)
               Presence of oily sheen
               Odor  (e.g., hydrogen sulfide,  oil, creosote)
     0    Gross characteristics of the  vertical profile
               Changes  in sediment characteristics
               Presence  and depth of  redox  potential discontinuity
               (rpd) layer


                                    12

-------
                                                                   Metals
                                                         Sample Collection
                                                              August 1986

     •    Penetration depth
     0    Comments related to sample  quality
              Leakage
              Winnowing
              Disturbance.

     Before subsamples of the surficial sediments are taken, the overlying
water must be removed.   The preferred method  of removing this water  is
by slowly siphoning it  off near one side of  the  sampler.  Methods such
as decanting  the water or slightly cracking the  grab to let the water run
out are  not  recommended, as  they may result  in unacceptable disturbance
or loss of fine-grained surficial  sediment and organic matter.

     Once the overlying water has  been  removed, the  surficial  sediment
can be subsampled.   It  is  recommended that  subsamples be taken  using a
flat  scoop  shaped like  a coal  shovel.  The shoulders of the scoop should
be 2 cm high.  This device will allow a relatively  large subsample  to  be
taken  accurately to a depth  of  2 cm.  Coring devices are  not recommended
because generally they  collect small amounts  of surficial sediment and
therefore require repeated extractions to  obtain a sufficient volume of
material  for  analysis of  conventional  sediment variables.  A curved  scoop
is not  recommended because it does not sample  a  uniform  depth.  Because
accurate and consistent  subsampling requires  practice, it is advisable
that an experienced person perform this  task.

     When subsampling surficial sediments,  unrepresentative  material should
be removed in the  field under the supervision of  the chief scientist and
noted on the  field log sheet.  The criteria used to determine representa-
tiveness should be determined prior to sampling.

     Finally, if  samples are  to  be analyzed for trace metals or  priority
pollutant organic  compounds, sample contamination during collection must
be avoided.   All  sampling equipment (i.e., siphon  hoses, scoops, containers)
should be made of noncontaminating material and should be cleaned appropriately
before use.   Samples  should  not be touched with ungloved fingers.   In addition,
potential airborne contamination (e.g.,  stack  gases, cigarette smoke) should
be avoided.   Detailed guidance for preventing sample contamination  is  given
in a previous section of  this chapter.

TISSUE SAMPLES

Collection

     The  major difficulty  in  trace metal  analyses  of tissue samples is
controlling contamination of the sample after collection.  In  the field,
sources  of contamination  include sampling  gear,  grease  from winches or
cables, engine exhaust, dust, or ice used  for  cooling.   Care must  be  taken
during handling to avoid  these and any other possible sources  of contamination.
For example,  during sampling the ship  should  be positioned such  that the
engine  exhausts do  not  fall  on deck.  To avoid  contamination  from melting


                                    13

-------
                                                                   Metals
                                                         Sample Col lection
                                                              August 1986

ice, the samples should be wrapped  in  aluminum foil  and placed  in watertight
plastic bags.  The outer skin of the fish or  shell  of the shellfish  acts
as protection  against metals contamination from the aluminum foil.

     Sample  resection and any subsampling of the organisms should be carried
out in a controlled environment (e.g., dust-free  room).  In most  cases,
this  requires that the organisms be transported  on ice to  a laboratory,
rather than  being resected on board the sampling vessel.  It is recommended
that  whole  organisms not  be  frozen prior to  resection if  analyses will
be conducted only on selected tissues, because freezing may cause  internal
organs to rupture and contaminate other tissue.  If organisms are eviscerated
on board the survey vessel,  the remaining tissue  (e.g., muscle)  may be
wrapped as described above and frozen.

     Resection  is best performed under "clean room"  conditions.  The "clean
room" should have positive  pressure and filtered  air.  The  "clean  room"
should  also be  entirely metal-free and isolated from all  samples high in
contaminants (e.g., hazardous waste).  At a minimum, care  should  be  taken
to avoid  contamination from  dust, instruments, and all  materials that may
contact the  samples.  The best equipment to use for trace metal  analyses
is made of  quartz, TFE, polypropylene, or polyethylene.   Stainless steel
that is resistant to corrosion may be used if necessary.  Corrosion-resistant
stainless steel is not magnetic, and thus can be distinguished from other
stainless steels with a magnet.  Stainless steel  scalpels  have been  found
not  to  contaminate mussel  samples (Stephenson et al. 1979).  However, low
concentrations of heavy metals in other biological tissues (e.g.,  fish
muscle)  may be  contaminated  significantly  by any exposure to  stainless
steel.  Quartz utensils are ideal but  expensive.  To control contamination
when resecting tissue, separate sets of utensils should be used for removing
outer tissue and for removing tissue  for analysis.   For bench liners and
bottles, borosilicate glass would be preferred over plastic if  trace organic
analyses are to  be performed on the same  sample.

     Resection should be conducted  by or  under the supervision  of a competent
biologist. For fish samples,  special  care must be  taken to avoid  contaminating
target  tissues  (especially  muscle) with slime and/or adhering sediment
from the fish  exterior  (skin) during resection.  The incision  "troughs"
are  subject to  such contamination.  Thus, they should not be included in
the sample.   In  the case of muscle, a "core" of  tissue is taken from within
the  area boarded by the incision  troughs, without contacting  them.  Unless
specifically sought as  a  sample,   the dark muscle  tissue  that  may exist
in  the  vicinity of the lateral line should  not  be mixed with  the light
muscle tissue  that consitutes the rest of the muscle tissue mass.

     Prior  to use,  utensils  and  bottles should be thoroughly cleaned with
a detergent  solution, rinsed with tap water, soaked in acid, and then rinsed
with  metal-free water.  For quartz, TFE, or glass containers,  use  1+1 HN03,
1+1  HC1,  or aqua regia  (3 parts  cone HC1 +  1  part cone HN03) for soaking.
For  plastic material, use  1+1 HNQ.3 or 1+1 HC1.   Reliable soaking  conditions
are  24  h at 70° C (Greenburg et   al. 1985).   Do  not use chromic acid for


                                    14

-------
                                                                   Metals
                                                        Sample Collection
                                                              August  1986

cleaning  any materials.  Acids  used should be at  least  reagent grade.
For metal  parts, clean as stated for glass or plastic,  except omit the
acid soak  step.  If trace organic analyses are  to  be performed on the  same
samples,  final rinsing with methylene chloride is acceptable.

     Sample size  requirements can vary with tissue type (e.g., liver  or
muscle)  and detection limit  requirements.   In  general,  a  minimum sample
size of  6  g (wet weight) is required for the analysis of all priority pollutant
metals.   To allow  for duplicates,  spikes, and required reanalysis, a sample
size of 50 g  (wet weight) is recommended.  Samples can be stored in glass,
TFE, or  high-strength polyethylene jars.

Processing

     Samples should be frozen after resection and  kept at -20° C.  Although
specific holding times  have  not been recommended  by U.S.  EPA, a maximum
holding time  of 6 mo  (except  for mercury samples,  which  should be  held
a maximum of  28 days)  would  be consistent with that for water samples.

     When  a sample  is  thawed, the associated  liquid should be maintained
as a part of the  sample.  This liquid will contain lipid material.   To
avoid loss of moisture from the sample, partially thawed samples should
be homogenized.  Homogenizers used to  grind the tissue should have tantalum
or titanium parts  rather than stainless steel parts.  Stainless steel blades
used during homogenization have been found  to  be  a source of nickel and
chromium contamination.
                                   15

-------
                                                                    Metals
                                                        Analytical  Methods
                                                               August  1986
                            ANALYTICAL METHODS
SAMPLE PREPARATION

Water:  Total  Metals

     Freshwater  samples to be analyzed by graphite furnace atomic absorption
spectroscopy  (GFAA)  should be prepared using  an HN03/H202 digest.  Analysis
by inductively-coupled plasma emission spectroscopy (ICP)  or direct flame
atomic  absorption  spectroscopy (DFAA) requires an HN03/HC1 digest.   Samples
to be analyzed for  mercury by cold vapor atomic absorption spectroscopy
(CVAA)  must have  a  separate,  100-mL aliquot digested using potassium  per-
manganate and potassium persulfate.   All  three procedures are detailed
in Exhibit  D  of the Contract  Laboratory  Program (U.S.  EPA 1985), which
is attached  to this report as Appendix B.

     Because  digestion creates excessive salt concentrations that interfere
with instrumental methods, saltwater samples must be extracted.  Extraction
with  ammonium pyrrolidine dithiocarbamate/methyl isobutylketone (APDC/MIBK)
is recommended and  is detailed in Appendix F (Greenburg et al. 1985).

Water;  Dissolved Metals

     According  to  convention,  "dissolved"  substances are  those contained
in the filtrate passing a 0.45 urn filter.  Samples to be analyzed for dissolved
metals  must be filtered  prior to preservation and subsequent digestion.
However, filters are  a significant source of trace element contamination.
Membrane filters  (e.g.,  polycarbonate,  fluorocarbon)  are significantly
cleaner than paper  (Murphy 1976).

      It  is  recommended that 0.45 urn polycarbonate  filters without  grid
marks be used  for the filtration of water samples and that they be  cleaned
by drawing  through  500 mL  1:1 HN03  followed  by 1 L metal-free water (to
neutral  pH).  Prior  to  use, filters  must be stored in  a tightly-closed
container to avoid  contamination from airborne dust.

     The filtrate should be acidified to pH<2  using 1:1 HN03.  If a precipitate
appears upon acidification, the sample must be digested as  for total metals.

Water;  Particulate Metals

     Particulate  substances are  operationally defined as those retained
on a 0.45 urn  filter.  Polycarbonate  filters should be  used and  cleaned
as detailed  in the  previous section.   After cleaning,  the filters should
be dried to constant weight at 60   C.  A detailed protocol for shipboard
filtration of large volumes of seawater using commercially-available equipment
is given by Mart (1979b).  A completely submersible, self-contained filtration
apparatus, capable of filtering up to 1,000 L at a  flow rate  of 50-200 mL/min,

                                    16

-------
                                                                    Metals
                                                        Analytical  Methods
                                                               August  1986

is commercially available  from Seastar  Instruments Ltd., Sidney,  British
Columbia,  Canada.  After  sample filtration,  the filters  should be dried
again to  constant weight at 60  C.  Total  sample weight is then calculated
by difference.

     After weighing, the entire filter plus  sample is digested as for  sediment
samples  as detailed in Appendix D.

Sediment;   Bulk  Analysis

     The two most  common digestion  methods employed for sediment  samples
are total  and  strong acid.  Total digestion  is the most severe  and employs
hydrofluoric acid to aid in the breakdown of silicate matrices and/or perchloric
acid to  aid in the decomposition of  organic  matter.  The use of a total
acid digest has been recently endorsed by the NOAA/NBS Quality Assurance
Workshop (December 5-6, 1985).

     Strong acid digestion employs  nitric acid,  some ratio of nitric and
hydrochloric acids, or  nitric acid  and  hydrogen  peroxide.   Strong  acid
digestions are not as efficient as total  acid methods for the solubilization
of certain metals (e.g., aluminum,  chromium) that are tightly matrix-bound,
but  they  do not require  special   labware  or  hoods.  The  efficiencies of
various  strong acid digestion methods  have  been reviewed  by Plumb  (1981).
Although Plumb concludes that the literature for  strong acid digests  supports
the  use of HN03:HC1, the U.S. Environmental  Protection  Agency Contract
Laboratory Program (CLP) requires  the use of  HNC^rHgOz (U.S. EPA 1985).
The detailed procedure for the latter  method is given in  Exhibit D of  the
CLP, which is  attached to  this report  as  Appendix B.

     A hydrofluoric  acid/aqua  regia total acid digestion was agreed  upon
at the January 17, 1986,  PSDDA/PSEP  workshop.  The detailed  protocol  is
attached to this report as Appendix D  (Rantala and Loring 1975).

Sediment:   Elutriate and Fractionation Studies

     The elutriate test is used to estimate  the types and amounts of substances
likely to be released to the water  column  from sediments  during dredging
operations.  Fractionation  studies  estimate the distribution and mobility
of substances  occurring in sediments. Detailed procedures for these  two
tests are given  by Plumb (1981) and are attached  to  this report as Appendix C.
The filtered samples  from the elutriate  test are treated as "dissolved
metals"  samples  discussed  above.  The  filtered  samples from the six different
"phases" of the  fractionation procedure should  be analyzed directly.

Tissue

     Workshop  participants agreed upon a  nitric acid/perchloric acid digestion
for tissue samples.  Perchloric acid is especially useful for the dissolution
of  fats.   The detailed procedure for  this  method is attached to this report
as Appendix E  (Tetra Tech  1986a).   Equivalent  methods may be used  (e.g.,


                                     17

-------
                                                                   Metals
                                                        Analytical Methods
                                                               August 1986

U.S. EPA  1977) if accuracy  and precision can be demonstrated  to  the level
specified  in  the Quality Assurance/Quality Control section of this  report.

INSTRUMENTAL  METHODS

     The instrumental methods currently  available for the analysis of trace
metals  include  electrochemical  (e.g., differential  pulse  polarography),
spectrophotometric (e.g., silver diethyldithiocarbamate), atomic absorption
spectrophotometric, atomic emission spectrophotometric, x-ray fluorescence,
and neutron activation.

     While any analytical method  might be used as long as the documented
instrument or method  detection limits  meet  the required  LODs specified
(Appendix  A),  certain  instrumental techniques have drawbacks and are not
recommended.   For example,  both electrochemical  and spectrophotometric
techniques are time-consuming and are not amenable to the analysis of large
numbers of  samples.  Neutron activation requires the use of expensive nuclear
equipment  that is not  widely  distributed.  Certain specialized techniques
designed for  the analysis of particular  metal  species (e.g.,  organotins)
also require the use of  research-grade instrumentation often extensively
modified from  commercially available equipment (e.g., Donard and  Weber 1985).

RECOMMENDED INSTRUMENTAL METHODS

     The LODs recommended in Appendix A were arrived at by matching U.S. EPA
water quality criteria (for water analyses)  and the lowest  concentrations
expected  in  Puget Sound sediments and marine  tissues against the  detection
limits  possible for each matrix using agency-approved instrumental  methods.

     For mercury,  the  only instrumental method  adequate for these requirements
for all matrices is  cold vapor atomic absorption  (CVAA).  For tissue, graphite
furnace atomic absorption  (GFAA)  is required  for all  the other priority
pollutant  metals of concern.

     For sediment,  inductively-coupled plasma  emission spectroscopy  (ICP)
methods are adequate  for copper,  nickel,  and zinc.  All  other  priority
pollutant metals of concern in sediment require GFAA to achieve the LODs.

     For water,  ICP  methods are adequate  for cadmium and zinc.  All other
priority pollutant metals of concern in water  require GFAA  to  achieve the
LODs.

     Atomic  absorption analysis via hydride generation (HYDAA)  can  lower
the detection limits  for antimony,  arsenic,  and selenium  by a  factor  of
10-100 (Skoog 1985).   This technique may  be  substituted for GFAA for  samples
in which low concentrations of  these three elements are expected, or  it
may  be  used  routinely.  The U.S. EPA-approved  method  for arsenic by  hydride
generation (U.S.  EPA  1979a) is included  in  Appendix  G and  can be adapted
for the analysis  of selenium and antimony.   Hydride generation may be coupled
to ICP instead of  AA  for certain  elements  (e.g., selenium  and antimony).


                                    18

-------
                                                                    Metals
                                                        Analytical Methods
                                                               August 1986

However,  this method  may be used only  if accuracy  and  precision can be
demonstrated to the  levels  specified in the Quality Assurance/Quality Control
section of this report.

     The U.S.  EPA  approved  methods for GFAA  and ICP analysis for all the
metals of concern except mercury and for the CVAA  analysis of mercury are
detailed  in Exhibit  D  of  Appendix B (U.S. EPA 1985).   It must be emphasized
that GFAA requires duplicate injections and matrix spikes.   If spike recovery
is not within specific  limits as detailed in the CLP (Exhibit E of Appendix B),
the method of standard  additions  (MSA) may be  required.   The monitoring
of GFAA spike recoveries is critical for all low level samples and  is especially
critical for the analysis  of  lead.

     The use  of matrix modifiers has become widespread in  the GFAA analysis
of freshwater (Manning  and Slavin  1983), saltwater  (Slavin et al.  1984,
Manning and Slavin  1978),  and biological samples (Hinderberger et al. 1981).
No agency-approved  protocol was  found.   However,  matrix modifiers  may be
used,  especially  for  lead,  arsenic,  selenium,  and  antimony, as long as
accuracy and precision  can be demonstrated to the  levels specified  in the
Quality Assurance/Quality  Control section of this report.

     Although  X-ray fluorescence  (XFS) may  achieve the LCDs  listed in
Appendix A, no agency-approved method was found.   XFS may be used if accuracy
and precision  can be  demonstrated  to the levels specified in the Quality
Assurance/Quality Control  section of this report.

     Other  agency-approved  instrumental  techniques  with detection limits
higher than those specified in Appendix A (e.g., direct flame atomic absorption)
may be used only  if the  sample concentration exceeds twice the documented
detection limit of  the  method.   These U.S. EPA-approved methods  are detailed
in Appendix G (U.S. EPA 1979a).

METALS SPECIATION

     Because the techniques for  the analysis of arsenic and organotin species
are not widely used,  are  not routine, and have inadequate QA/QC documentation,
no protocols were recommended at the January PSDDA/PSEP workshop.  Although
not yet tested,  it was  suggested that methyl mercury may possibly  be  extracted
by vacuum distillation  (e.g., Hiatt  and Jones 1984) and then analyzed by
gas chromatography/mass spectrometry.
                                    19

-------
                                                                    Metals
                                                                     QA/QC
                                                               August 1986
                    QUALITY ASSURANCE/QUALITY CONTROL
QA/QC MEASURES  INITIATED BY THE ANALYTICAL LABORATORY

     Standard  laboratory practices for cleanliness as applied to glassware,
reagents,  solvents, gases, and instruments must  be followed.  For additional
guidelines  not covered  in  this report,  see  Sections 4 and 5 of Handbook
for Analytical Quality Control in Water and Wastewater Laboratories (U.S.  EPA
1979b).

Instrumental  QA/QC Checks

     Instrumental  QA/QC checks  necessary for  all  the U.S. EPA-approved
methods discussed  in the previous section include:

     •    Calibration blank

     •    Initial  calibration and calibration verification

     •    Continuing calibration verification

     •    U.S.  EPA Standard Reference Samples

     •    ICP interference check sample analysis (for ICP only).

     Guidelines  for instrumental calibration  are given in U.S. EPA Methods
for Chemical  Analysis of Water and Wastes (U.S. EPA 1979a).  In general,
all  instruments must  be  calibrated  daily  and each time the instrument is
set up.  Calibration procedures will  follow the procedures  specified  for
each  analysis in the  U.S.  EPA protocols.   In addition (as specified for
the CLP, Appendix  B), after an instrument system has been  calibrated,  the
accuracy  of the initial  calibration shall  be verified and documented for
every analyte by  the analysis of U.S. EPA Quality Control Solutions.  Where
a  certified solution  of  an analyte  is not  available  from U.S. EPA or any
source, analyses  shall  be conducted on an independent standard at a concen-
tration other than that  used for calibration, but within the calibration
range (e.g.,  tin is not currently present in U.S.  EPA Quality Control Samples).
When  measurements for the  certified components exceed the control limits,
the analysis must be  terminated, the problem  corrected,  the instrument
recalibrated, and  the calibration reverified.

     For  ICP  and  AA analyses, all work shall  be performed using continuing
calibration as outlined  in  the CLP Statement of Work with  regard to Exhibit E,
Sections  1  and 2 (see Appendix B).   Frequency is  10  percent or every 2 h
during an analysis run, whichever is more frequent.
                                    20

-------
                                                                    Metals
                                                                     QA/QC
                                                               August 1986
Method QA/QC  Checks
     Contract  laboratories should perform  the quality control  checks  listed
below:

     0     Preparation blank

     •     Spiked  sample analysis

     t     Replicate  sample analysis

     0     GFAA method of standard addition  (if necessary)

     0     Laboratory control sample analysis  (Standard Reference Materials).

     Details  on  the use and application of  these checks along with control
limits, reporting requirements, and corrective actions  are  discussed  in
Exhibit  E  of Appendix B  (U.S. EPA 1985).  The frequencies of application
of these checks are  5 percent or one per batch,  whichever is more frequent.
The control  limits are +20 percent relative percent difference  for duplicates,
75-125  percent recovery  for spikes, and 80-120 percent  recovery for the
analysis  of standard reference materials (SRMs).  For the purpose of  QA/QC,
the contract required detection limits  (CRDLs) shall  be  those listed  in
Table  6  of  Appendix A.   For  batches of  five  samples or less, the minimum
QC checks should  be  a preparation  blank and the analysis  of an SRM.   If
an analyte  is not  in the SRM, a matrix  spike must be analyzed for that
particular analyte.

     In general  for small batches (i.e.,  analytical duplicates  >  matrix spikes.   If
several  small batches of the same matrix are  analyzed sequentially  (i.e.,
for several  small projects),  an SRM can be analyzed at a  frequency of 5 percent
overall,  with at least one sample duplicate analyzed per individual  batch.
If any QA/QC check does not meet the established criteria,  the laboratory
QA officer  should  notify the  project QA coordinator  and the methods  should
be adjusted or,  if necessary, data qualifications and reasons for noncompliance
to criteria should be submitted with the analytical data report.

QA/QC MEASURES INITIATED IN THE FIELD

     In addition  to  the QA/QC checks listed above, the following four  checks
may be initiated at the time  of  sample collection  (Plumb 1981).   These
checks may not  replace any of the QA/QC measures outlined above but may
be included as part  of the overall QA/QC program.
                                    21

-------
                                                                    Metals
                                                                     QA/QC
                                                               August 1986
Transfer (Preservation) Blanks
     Reagents can  become contaminated  after a period of use in the  field.
The transfer  blank will shown this and any  contamination  introduced during
shipping.

     A sample container is filled  with distilled water to the same  volume
as that for  samples and preserved as  if it  were a normal  water or sediment
sample. This blank is then sent to the laboratory for analysis.

Cross-Contamination Blanks

     Field equipment can exhibit carryover from one  sample  to  the  next
if not thoroughly cleaned between sampling.   The cross-contamination blank
is designed  to verify the absence of  carryover.

     Decontaminated sample-handling equipment (spatulas, augers, core barrels)
is wiped with  a clean lab tissue, which  is then  placed in a scalable container.
Alternatively,  the  equipment is rinsed  with  distilled/deionized  water,
and the water  is  collected and preserved as  if it were a normal water sample.

Blind Replicate  Samples

     A collected sample is homogenized and split in the field into at least
three identical  aliquots,  and each  aliquot  is  treated  and  identified  as
a  separate  sample.  The  replicates are sent blind to the laboratory.   The
mean,  standard deviation, and relative percent  standard deviation are calculated
by the project QA coordinator.

     In addition,  a  collected  sample may  be split  in the field into two
aliquots,  and one aliquot sent for analysis to a different  or "reference"
laboratory.   The relative  percent  difference is calculated by the project
QA coordinator.   If project constraints  require the use  of  more than  one
laboratory,  their comparability must  be established using certified reference
materials.

Blind Standard Reference Materials

     A standard reference material is placed in a sample container  at the
time of sample collection and sent blind to the laboratory.  The percent
recovery is  calculated by the project QA coordinator.

     It is recommended that  the same control limits be established  as for
the laboratory QA/QC  checks.  The project  QA coordinator must inform  the
laboratory  if the control  limits are exceeded and corrective actions must
be taken.
                                    22

-------
                                                                   Metals
                                                                    QA/QC
                                                               August 1986
CORRECTIVE  ACTIONS
     If the concentration of the field or laboratory blank is  greater than
the required detection limit, all  steps in the  sample handling  should be
reviewed.  Many trace metal  contamination problems are due to airborne
dust.  Keeping containers closed, and rinsing all handling equipment immediately
before use  minimizes dust problems.   In the field, mercury-filled thermometers
are a potential  source of severe mercury contamination.  In the  laboratory,
samples for mercury analysis should  be  isolated from items such as polarographs
or COD reagents.

     Poor  replication may be caused  by  inadequate  mixing of the sample
before taking aliquots, inconsistent contamination, inconsistent  digestion
procedures,  or instrumentation  problems.  Instrumentation  problems may
be isolated by  recalibration and calibration blank analysis.   Inconsistent
digestion  may  be due to  analyte loss; see below for corrective action.
Also, hotplates may not hold a constant temperature across  their  surfaces.
This can be remedied by changing the position of digestion vessels at regular
intervals during heating.

     Poor  performance on the analysis  of the Standard Reference Material
(SRM) or poor spike recovery may be caused  for  the  same  reasons  as poor
replication.  However, if replicate results are acceptable, poor SRM  performance
or poor spike recovery may be caused by  loss  of  analyte during digestion.
To check for  analyte loss  during digestion and for low  recovery due to
interferences during analysis, spike the  sample after digestion  and  compare
the  analysis  to the  predigestion spike.   If the results are different,
the digestion technique should be adjusted.   If  the results are  the same,
dilute  the sample by at  least a factor of five and reanalyze.   If  spike
recovery is still  poor, standard additions,  matrix modifiers, or  another
method is required.
                                    23

-------
                                                                    Metals
                                                            Data Reporting
                                                               August 1986
                              DATA  REPORTING


     Concentrations of elements  from sediment  samples should be reported
on a dry weight  basis.   Those from  tissue samples  should be reported on
a wet weight basis along with the percent moisture  content (wet/dry ratio)
of the tissue.  If the tissue sample is too small  to do both  a metals analysis
and a moisture determination, omit the latter.

DATA REPORT PACKAGE

     The data report package for analyses of each  sample should include:

     •    Tabulated results  in  units as  specified for  each  matrix
          in the analytical protocols (Appendices  B, D, E,  and G),
          validated  and  signed in original by the  laboratory manager

     0    Any data  qualifications  and  explanation for  any variance
          from the analytical protocols

     •    Results for all of the QA/QC checks initiated by the labora-
          tory

     •    Tabulation of  instrument and method detection  limits.

BACKUP DOCUMENTATION

     All contract laboratories  are required  to submit  results that are
supported by sufficient  backup data and quality assurance  results  to  enable
independent QA reviewers to conclusively determine the quality of  the data.

     The laboratories  should be able to supply legible photocopies  of original
data sheets with sufficient information to unequivocally identify:

     t    Calibration  results

     •    Calibration  and preparation blanks

     0    Samples and  dilutions

     •    Duplicates and spikes

     •    Any anomalies  in instrument performance or  unusual instrumental
          adjustments.
                                     24

-------
                                                                    Metals
                                                                References
                                                               August 1986
                                REFERENCES
American Society  for  Testing and Materials.  1983.  Annual  book of ASTM
standards.   Water and Environmental  Technology.  Vol. 11.01.  ASTM,  Phila-
delphia, PA.   752 pp.

Bates,  T.S., S.E.  Hamilton, and J.D.  Cline.   1983.  Collection of suspended
particulate matter  for hydrocarbon analyses:   Continuous flow centrifugation
vs. filtration.   Estuarine Coast.  Shelf Sci.  16:107.

Donard, O.F.X., and J.H.  Weber.  1985.   Behavior of methyltin compounds
under simulated estuarine conditions.   Environ.  Sci. Technol. 19:1104-1110.

Greenberg, A.E.,  R. Trussel, and L.S.  Clersceri (eds).   1985.  Standard
methods for  examination  of  waste and  wastewater.  American Public Health
Association,  Washington, DC.  1268 pp.

Hiatt,  M.H., and  T.L. Jones.   1984.  Isolation  of purgeable organics from
solid matrices by  vacuum distillation.   U.S.  Environmental Protection Agency,
Las Vegas Laboratory, Las Vegas, NV.

Hinderberger, E.J., M.L.  Kaiser, and  S.R. Koirtyohann.   1981.  Furnace
atomic absorption analysis of biological samples  using  the L'Vov platform
and matrix modification.  Atomic Spec. 2:1-7.

Horowitz, A.J.  1986. Comparison of methods for the  concentration of suspended
sediment in  river water  for subsequent chemical  analysis.  Environ.  Sci. &
Technol.  20:155-160.

Manning, D.C.,  and W. Slavin.  1978.   Determinatin of lead in a chloride
matrix with  the graphite furance.  Anal. Chem. 50:1234-1238.

Manning, D.C.,  and W. Slavin.  1983.  The determination of trace elements
in natural waters  using the  stabilized temperature  platform  furnace.  Applied
Spec. 37:1-11.

Mart,  L.  1979a.  Prevention  of contamination and other  accuracy risks
in voltamnetric trace metal  analysis of natural waters.   Part  I:  Preparatory
steps,  filtration,  and storage of water samples.  Fresenius Z. Anal. Chem.
296:350-357.

Mart,  L.  1979b.  Prevention  of contamination and other  accuracy risks
in volammetric trace metal  analysis of natural waters.   Part  II:  Collection
of surface water  samples.   Fresenius  Z. Anal. Chem.  299:97-102.
                                    25

-------
                                                                    Metals
                                                                References
                                                               August 1986
Murphy,  T.J.   1976.  The  role of the analytical  blank in accurate trace
analysis,  pp.  551-539.   In:  Accuracy in  Trace  Analysis:  Sampling,  Sample
Handling,  and Analysis.   National Bureau of Standards Special Publication
422.  NBS,  Washington, DC.

Plumb, R.H.,  Jr.  1981.   Procedure for  handling  and chemical analysis of
sediment  and water  samples.  Technical  Report EPA/CE-81-1.  U.S.  EPA  and
Corps of  Engineers,  U.S. Army Engineers Waterways Experiment Station, Vicksburg,
MS.

Rantala,  R.T.T.,  and  D.H. Loring.  1975.  Multi-element analysis of silicate
rocks and marine  sediments  by atomic absorption  spectrophotometry.   Atomic
Abs. Newsletter.  14(5):117-120.

Skoog, D.A.  1985.  Principles of instrumental analysis.  Third edition.
Saunders, Philadelphia, PA.  pp. 270-277,  282-284, 303-304.

Slavin,  W., G.R. Carnrick, and D.C. Manning.  1984.   Chloride  interferences
in graphite furnace atomic  absorption spectrometry.  Anal. Chem. 56:163-168.

Stephenson, M.D., M. Martin, S.E. Lange, A.R. Flegal,  and J.H. Martin.
1979.  California mussel watch  1977-1978.   Volume II:  Trace metals  concen-
trations in the California mussel, Mytilus californianus.  SWRCB Water
Quality Monitoring  Report No. 79-22.  Sacramento, CA.  110 pp.

Tetra Tech.  1985.   Commencement Bay nearshore/tideflats remedial investiga-
tion.  Final Report (Volume I).   Prepared  for Washington Department of
Ecology and U.S.  EPA. Tetra Tech,  Inc., Bellevue, WA.

Tetra Tech.  1986a.  Bioaccumulation monitoring  guidance:  4.  analytical
methods  for EPA priority pollutants and 301(h) pesticides in tissues  from
estuarine and  marine organisms.   Final report  prepared  for  the  U.S.  EPA
by Tetra Tech, Inc., Bellevue,  WA.
Tetra  Tech.  1986b.   Protocols for  collection of particulate  samples
organic and inorganic analysis.  Tetra Tech, Inc., Bellevue, WA.
for
U.S. Environmental  Protection Agency.  1977 (revised October,  1980).  Interim
methods for the sampling and  analysis of priority pollutants  in sediments and
fish tissue.  Environmental  Monitoring and Support Laboratory, Cincinnati, OH.

U.S. Environmental  Protection Agency.  1978 (revised 1983).   NEIC  Policies
and  procedures.   EPA-330/9-78-001-R.  National Enforcement  Investigations
Center, Denver, CO.

U.S. Environmental  Protection Agency.  1979a (revised March,  1983).  Methods
for chemical analysis of water  and wastes.  EPA-600/4-79-020.   Environmental
Monitoring and Support Laboratory, Cincinnati, OH.
                                    26

-------
                                                                    Metals
                                                                References
                                                               August  1986
U.S. Environmental Protection Agency.  1979b.   Handbook for  analytical
quality control  in  water and wastewater  laboratories.  EPA-600/4-79-019.
National  Environmental Research Center,  Cincinnati, OH.

U.S. Environmental  Protection Agency.   1982.   Handbook  for  sampling and
sample  preservation of water and wastewater.  EPA-600-82-029.  Environmental
Monitoring and  Support Laboratory, Cincinnati, OH.  402 pp.

U.S. Environmental Protection Agency.   1984.  Guidelines establishing  test
procedures for the analysis  of pollutants.  U.S.  EPA, Washington,  DC.
Federal Register, Vol. 49, No. 209, pp.  43234-43436.

U.S. Environmental  Protection Agency.   1985.   Contract laboratory program
statement of work (SOW), inorganic analysis, multi-media,  multi-concentration.
SOW No. 785. U.S.  EPA, Washington, DC.
                                    27

-------
                 APPENDIX A

                FINAL REPORT
                  TASK A-l

SELECTION OF METALS FOR PROTOCOL DEVELOPMENT

-------
                                 CONTENTS





                                                                        Page




LIST OF TABLES                                                        A-1ii



TASK A-l:  SELECTION OF METALS FOR PROTOCOL DEVELOPMENT                 A-l




    INTRODUCTION                                                        A"1



    EVALUATION OF U.S. EPA PRIORITY POLLUTANT METALS                    A-2




    METALS SPECIATION                                                   A~5



    EVALUATION OF NON-PRIORITY POLLUTANT METALS                         A-5



    RECOMMENDED LIMITS OF DETECTION                                     A-6




    SUMMARY                                                            A'14



REFERENCES                                                             A~15
                                     A-li

-------
                                  TABLES


Number                                                                  Page

  1     Metals of concern as indicated by Puget Sound  field  studies      A-3

  2     Limits of detection for water, sediment, and tissue
        matrices by instrument                                          A-9

  3     U.S.  EPA Water Quality Criteria                               A-10

  4     Summary of metal concentrations in sediments from Puget  Sound
        reference areas                                                A-ll

  5     Minimum and maximum metal  detection limits reported  for
        tissue samples                                                 A-12

  6     Recommended limits of detection for water, sediment,
        and tissue matrices                                            A-13
                                   A-iii

-------
                                TASK A-l
               SELECTION OF METALS FOR PROTOCOL DEVELOPMENT
INTRODUCTION

     Recent studies  have documented high  concentrations of  a  number of
U.S. EPA  priority pollutant metals  and metalloids  in  the  sediments  of
industrialized embayments  in  Puget  Sound relative  to  reference areas.
To ensure  the  acceptability of dredging and dredged material disposal,
federal and state agencies must be able to accurately  and routinely analyze
for these  substances in Mater, sediment, and biological samples.  Because
of the number of agencies  involved  and different program  objectives, a
standardized  protocol is  necessary to ensure comparability of results.
It is the  objective of Task A to develop this protocol.

     A consensus  approach involving  federal,  state, and  local institutions
will be used to develop both standardized chemical  tests  and  standardized
test interpretation guidelines for use by all Puget Sound  regulatory agencies
concerned  with dredged materials. Protocols developed for each metal  (or
group  of  metals  that can be analyzed simultaneously)  will  include guidance
for field  sampling, analytical procedures, and quality  control requirements.
When possible, existing  procedures  are to be  recommended, and modified
as appropriate.   Draft protocols will  be reviewed by a  workshop of experts
to  reach  a consensus on the recommended requirements. After incorporation
of the workshop comments, a final protocol report will  be  drafted  for  final
review.

     The purpose  of Task A-l is to:

     •    Recommend appropriate metals and metalloids for which analytical
          protocols  will be  proposed under the Puget Sound  Dredged
          Disposal Analysis (PSDDA)  program
                                   A-l

-------
     •    Recommend  limits of detection (LOD) for the  analysis of
          each element in water,  sediment, and tissue samples.

     The thirteen U.S. EPA priority  pollutant metals include three elements
(antimony,  arsenic, and selenium)  that are classified as metalloids,  which
are  elements that do not  strictly occur as metals in the environment.
Following U.S.  EPA convention and for ease of discussion, all  of  these
elements will be  referred to as metals.

EVALUATION OF U.S. EPA PRIORITY POLLUTANT METALS

     Protocols were  requested by PSDDA  for all metals  either documented
or likely to be of concern in Puget Sound.   In addition, justification
was required for  the exclusion of  any U.S. EPA priority  pollutant metal.

     Reports for  four recent field investigations in Puget  Sound  were reviewed
during the selection process (Malins  et al. 1980, 1982;  Galvin et al.  1984;
Battelle 1985; Tetra Tech 1985a).   Historical data from previous  investigations
summarized by Dexter et al. (1981) and recommendations for  metals of concern
proposed  in several  summary  reports  of contamination in Puget Sound were
also reviewed (Konasewich et al.  1982;  Jones and Stokes 1983; Quinlan et
al. 1985).

     The thirteen U.S. EPA priority pollutant metals and data reports providing
evidence of their potential concern in Puget  Sound are listed  in  Table 1.
Criteria  typically used to determine if a metal is of  concern  include high
toxicity (measured in laboratory studies), high concentration  relative
to sediments from reference  areas  located away from  industrialized areas
of Puget Sound, and widespread distribution in Puget Sound.  While all  U.S. EPA
priority pollutant metals are considered toxic, metals identified  to be
of concern have  been found  in  high concentration only in limited areas
of the Sound (e.g., antimony, arsenic, cadmium,  copper, lead,  mercury,
and  zinc  at the  Fourmile Rock disposal site relative to the rest of central
Puget Sound; Galvin et al. 1984).
                                  A-2

-------
   TABLE  1.   METALS  OF  CONCERN  AS  INDICATED  BY  PUGET  SOUND  FIELD  STUDIES
U.S. EPA
Priority Dexter Mai ins Galvin
Pollutant et. al . et. al . Battelle et. al.
Metal 1981* 1982 1985 1984
Antimony (Sb) x x
Arsenic (As) x x x x
Beryllium (Be)
Cadmium (Cd) x x
Chromium (Cr)
Copper (Cu) x x x x
Lead (Pb) x x x x
Mercury (Hg) x x x x
Nickel (Ni)
Selenium (Se) x
Silver (Ag) x x
Thallium (Tl)
Zinc (Zn) x x x x
Tetra
Tech
1985a
x
x

X

X
X
X
X

X

X
Frequency
of Concern
(Out of 5)
3
5
0
3
0
5
5
5
1
1
3
0
5
* Secondary source (summarizes pre-1980 historical data).
                                    A-3

-------
     Two areas of  consensus are  indicated by Table 1.   First, three metals
(beryllium,  chromium, and thallium)  were not  identified by any report  as
of concern.   Although potentially toxic, these metals are not present in
Puget Sound  sediments in concentrations that exceed natural levels.   These
three metals  are not recommended  for protocol  development.  Second, five
metals (arsenic,  copper, lead, mercury, and zinc)  were identified  in  all
reports  as of  concern because of their high concentration in urban embayments.
A potential  association with adverse biological  effects was also noted
in some  reports.  These five metals  are recommended for  protocol development.

     Three additional  metals  (silver,  cadmium, and antimony) were found
in elevated  concentration  by a majority of the  investigators.  Of these
metals,  silver was found  in high  concentrations  at  relatively few sites
(Malins  et al.  1982; Battelle 1985;  Tetra Tech 1985a).   Silver  is extremely
toxic even  at low concentrations, it is a  potentially useful tracer of
sewage effluent,  and historically  its concentration in  Puget Sound sediments
has  increased faster than  that  of other metal  pollutants (Crecelius,  E.,
22 October 1985,  personal communication).  Because of these factors,  silver
has  been identified as a  metal  of concern.  Cadmium is also highly toxic
and is recommended  for  protocol development.  Although antimony  is less
toxic, its high concentration in selected areas (e.g.,  sediments contaminated
with copper  smelter slag) warrants its  inclusion.

     Selenium  was  reported in high concentration in only one  investigation
(Malins  et  al. 1982).   Several summary reports subsequently used these
data  to  justify including  selenium as a metal  of concern  (Konasewich et
al. 1982; Jones and Stokes 1983; Quinlan et al.  1985).  The  instrumental
technique used to  measure selenium  in  the  investigation (ion-coupled  plasma
emission spectroscopy)  is  subject to substantial  spectral interferences
(Parson et al.  1980).  All other investigations using alternative techniques
(e.g., graphite  furnace  atomic adsorption)  have reported uniformly  low
or undetected  concentrations of selenium.   Therefore, selenium  is not  recom-
mended for protocol development.

     In Commencement Bay, nickel was found  in  somewhat elevated concentrations
at a few sites exhibiting biological effects (Tetra Tech  1985a).  Although

                                   A-4

-------
the available  data  suggest that nickel  primarily derives from natural sources,
there are insufficient data to dismiss nickel as a potential  metal  of concern
in those few areas.  Therefore, nickel  is  recommended for protocol development.

METALS SPECIAT ION

     The toxicity  of  metals  is  highly dependent upon their  chemical  form
(speciation).   For  example,  inorganic  arsenic in the +3  oxidation  state
is much  more  toxic than  in  the +5 oxidation state.  Inorganic arsenic  is
toxic if ingested by humans, yet organic arsenic  compounds are relatively
nontoxic to mammals.   In contrast, mercury  salts are relatively nontoxic
because of low solubilities, but methylated mercury compounds are  highly
toxic.   Hence,  quantitative determinations of different metal  species could
seriously affect interpretation of dredged material test results.

     Recent studies (e.g., Crecelius et al. 1984; U.S. EPA 1980)  have shown
that arsenic and mercury exist predominantly  as methylated  compounds  in
marine biota.   Dominance by one chemical form reduces the need for quantitative
speciation studies  as routine measurements.  However, in some  cases  (e.g.,
health risk assessment) a precise determination of metals species in different
bioaccumulation  studies may be necessary.

     In the water  column, total  concentrations of arsenic and mercury are
low because of  dilution, and in sediments,  matrix effects confound speciation
studies.   Speciation techniques in general are difficult, time consuming,
and nonroutine  (Crecelius et al. 1984). Because of the apparent  small
incremental gain in knowledge from speciation analyses in  biota for all
but a few applications, routine analyses do not appear to be cost effective.
Development of procedures  for speciation analyses to be used in special
circumstances  will  require a consensus  from a  workshop of experts,  to  be
held as part of  Task A.

EVALUATION OF  NON-PRIORITY POLLUTANT METALS

     None of  the  reports reviewed  included a metal  of  concern that was
not already on the  U.S.  EPA priority  pollutant  list.  However, selected

                                   A-5

-------
non-priority pollutant metals could be of use  in  studies of dredged material.
Their value in  determining sediment quality values is being  investigated
as part of a separate task for PSDDA and the Puget Sound Estuarine Program.

     Chemical data can be normalized to account for physical and physicochemical
differences among samples.  These differences  may mask trends  in chemical
concentrations that are useful  in interpreting  environmental  data.  Metals
from pollution  sources tend to be selectively  enriched within  fine-grained
sediment fractions having a high content of  hydrous manganese or iron oxides.
The incorporation of metals in these oxides  may also affect their toxicity.
Therefore, normalization of metal concentrations to the iron or manganese
content of sediments may be useful for determining sediment quality values.
It may  also help account for the dilution  of  contaminated  sediments  by
materials from  unrelated sources during transport and deposition.

     Manganese  and iron are recommended for  inclusion in the metals protocol.

RECOMMENDED LIMITS OF DETECTION

     Environmental  analytical  chemists have not universally agreed upon
a convention for determining and  reporting  the lower detection limits  of
analytical procedures.  Furthermore, the basis for detection limits reported
in the literature is  rarely given.   Values  reported as  lower detection
limits  are commonly based  on instrumental  sensitivity,  levels of blank
contamination, and/or matrix interferences and  have various levels of statis-
tical significance.  The American Chemical Society's Committee on Environmental
Improvement (CEI) defined  the following types of detection  limits  in  an
effort to standardize the reporting procedures of environmental  laboratories
(Keith et al. 1983):

     •    Instrument Detection Limit  (IDL)  — the  smallest  signal
          above background noise that an instrument can detect reliably.

     •    Limit of Detection (LOD)  — the lowest concentration level
          that  can be determined to be  statistically different from
                                    A-6

-------
          the blank.   The recommended value for LOD  is  3o, where o
          is  the  standard deviation of  the blank in replicate analyses.

     •    Limit of Quantitation (LOQ) — the level above which quantitative
          results  may be obtained with a specified degree of  confidence.
          The recommended value for LOQ is 10o, where o is the standard
          deviation of blanks in replicate analyses.

     •    Method  Detection  Limit (MDL) —  the  minimum concentration
          of  a substance that can be identified, measured, and reported
          with 99 percent confidence  that  the  analyte concentration
          is  greater than zero.  The MDL is  determined from  seven
          replicate analyses  of a sample of a given matrix containing
          the analyte (Glaser et al. 1981).

The CEI  recommended that results below  3o should be reported  as "not detected"
(ND) and that the detection  limit  (or  LOD)  be given in parentheses.  In
addition,  if the results are near the detection limit (3o to 10a, which
is the "region of  less-certain quantitation"), the results should  be reported
as detections with the limit of detection given in parentheses.

     The CEI  definitions are useful for establishing a conceptual framework
for detection limits, but are somewhat  limited  in a practical  sense.  The
IDL does  not address possible blank  contaminants or matrix interferences
and is not a  good standard  for complex environmental matrices,  such as
tissues.   The LOD and LOQ  account for blank contamination,  but  not for
matrix complexity and  interferences.   The  high 10o level  specified for
LOQ helps  to preclude false  positive  findings,  but may also  necessitate
the rejection of  valid  data.  The MDL  is the  only operationally  defined
detection  limit  and provides a high  statistical confidence level but, like
the LOQ, may  be too stringent and necessitate the rejection of valid  data.

     The detection limits recommended  in this  report are not strictly  based
on the CEI definitions.  Instead,  they are  considered to be typically attainable
values  based on  the best professional  judgment and experience of analytical
chemists who considered the instrumental  sensitivity of affordable equipment,

                                   A-7

-------
common problems with blank contamination and matrix  interferences, and
reasonable  levels of laboratory analytical effort.   The recommended  values
are not  absolute,  as analytical  procedures and  laboratory precision can
affect attainable detection levels.  The detection  limits  recommended herein
fall  between  the IDL and MDL as defined by the CEI.

     The LOD for metals selected  for protocol development must be specified
low enough  so that data  from the  analysis of a particular matrix  (i.e.,
water, sediment,  or biota) will  be  comparable and of acceptable quality.
LOD reported  for the recommended metals are summarized in Table 2 for  each
matrix type  analyzed by a variety of instrumental  techniques:  direct flame
AA (DFAA),  graphite furnace AA  (GFAA), cold vapor AA (CVAA), and ion-coupled
plasma emission  spectroscopy (ICP).

     Several  factors  affect achievable detection  limits regardless of the
analytical  procedure.  These include the available sample  size, presence
of interfering  substances, range of  pollutants  to be analyzed, and level
of pollutant introduced as a contaminant during  sampling  or analysis.
LOD given  in Table 2 are provided as examples of typical  attainable levels.
These limits  are not intrinsically constant  for a  particular method,  but
also depend  on  the precision  attainable by an individual laboratory.  Each
laboratory must evaluate its own precision and thus establish its own detection
limits.  Large  variations from  the  LOD in Table  2 may  occur depending on
the analytical instrumentation and  technique and on  individual sample charac-
teristics.

     Recommendations of LOD must be based, in part,  on the analytical  require-
ments to  quantify concentrations found in  relatively "clean" reference
areas, and to  approach water criteria levels.  Table 3  lists the U.S. EPA
water quality criteria.  The  range of concentrations of the recommended
metals in  sediments from  Puget  Sound reference areas is shown in Table 4.
The range of  detection limits reported for these metals in marine  tissue
samples  is shown  in Table 5.  Recommended LOD are summarized in Table 6.
The recommended  LOD for sediments  and tissues were selected  by considering
attainable LOD  for different  instruments (Table 2), U.S. EPA water quality
criteria  (Table  3), and  reported  environmental  levels (Tables 4  and 5).
                                   A-8

-------
            TABLE 2.   LIMITS OF DETECTION FOR WATER, SEDIMENT,
                    AND TISSUE MATRICES  BY  INSTRUMENT


Antimony
Arsenic
Cadmium
Copper
Iron
Lead
Mercury
Manganese
Nickel
Silver
Zinc

(Sb)
(As)
(Cd)
(Cu)
(Fe)
(Pb)
(Hg)
(Mn)
(Ni)
(Ag)
(Zn)
ICP
0.032
0.053
0.004
0.006
NAd
0.042
0
NA
0.015
0.002
0.007
Water3
DFAA
0

0
0

0
.2
—
.005
.02
NA
.1
GFAA
0.
0.
0.
0.

0.
003
001
0001
001
NA
001
.0002 (CVAA)

NA
0.04
0
0
.01
.005

0.
0.
0.
NA
001
0002
00005
Sedimentb
ICP GFAA
3.2
—
4.0
0.6
0.7
4.2
0.01
2.0
1.5
0.7
0.2
0.1
0.1
0.1
0.1
—
0.1
(CVAA)
—
0.1
0.1
0.2
ICP
10
3
0.4
0.6
NA
4
0
NA
1
0.7
0.2
Tissue0
DFAA GFAA
—
—
0.1
0.1
NA
1.0
0.02
0.02
0.01
0.01
NA
0.03
.01 (CVAA)
NA
0.5
0.1
0.1
NA
0.02
0.01
0.2
a DFAA  and GFAA data  from U.S. EPA 1979;  ICP  data from U.S.  EPA 1984.
Values are mg/L.

b ICP data from Tetra Tech  1984;  GFAA and CVAA data are detection  limits
that can be reasonably attained by various  laboratories.   Under strict
conditions these limits  can be lowered (e.g., Battelle 1985).   Values are
mg/kg dry weight  for 5 g (wet) sediment  in a 100 mL digest.

c Tetra  Tech 1985b.  Values are ug/g  wet weight for 5 g tissue  in a 50 mL
digest.

d NA = Not applicable;  iron and manganese used as natural tracers for sediments
only.
                                  A-9

-------
TABLE 3.  U.S. EPA WATER QUALITY CRITERIAa (ug/L)
Metal
Antimony
Arsenic
Cadmium
Copper
Lead
Mercury
Nickel
Silver
Zinc
EPA UQC
—
36
9.3
2.9
5.6
0.025
7.1
2.3
58
a U.S. EPA  1980a and U.S. EPA  1985.
                        A-10

-------
          TABLE 4.  SUMMARY OF METAL CONCENTRATIONS  IN SEDIMENTS
                     FROM PUGET SOUND REFERENCE AREAS
Range
(mg/kg dry wt)
Antimony
Arsenic
Cadmium
Copper
Lead
Mercury
Nickel
Silver
Zinc
U O.lb
1.9 -
0.1 -
5 -
U 0.1
0.01 -
4 -
0.02 -
15 -
- 1.7
17
1.9
74
- 24
0.28
47
3.3
100
Mean
(mg/kg dry wt)
0.32C - 0.38d
7.2
0.67
32
9.8C - 9.8d
0.08
28
1.2
62
Detection
Frequency
12/32
34/34
24/24
28/28
21/28
38/38
26/26
26/26
26/26
Reference
Sites*
1,2,3,4,7,8,9
1,2,3,4,7,8,9
1,2,3,4,6,9
1,2,3,4,5,6,9
1,2,3,4,5,6,9
1-9
1,2,3,4,5,9
1,2.3,4,5,9
1,2,3,4,5,9
a Reference sites:  1.  Carr Inlet   4.  Case Inlet    7.  Nisqually Delta
                    2.  Samish Bay   5.  Port Madison  8.  Hood Canal
                    3.  Dabob Bay    6.  Port Susan    9.  Sequim Bay

b U:   Undetected at the limit of detection shown.

c Mean calculated using 0.00 for undetected values.

d Mean calculated using the reported detection limit for undetected values.


References:

      (Site 1)  Tetra Tech 1985a; Crecelius et al.  (1975).
      (Site 2)  Battelle (1985).
      (Site 3)  Battelle (1985).
      (Site 4)  Crecelius et al. (1975); Mai ins et  al. (1980).
      (Site 5)  Mai ins et al. (1980).
      (Site 6)  Mai ins et al. (1981).
      (Site 7)  Crecelius et al. (1975).
      (Site 8)  Crecelius et al. (1975).
      (Site 9)  Battelle (1985).
                                   A-ll

-------
TABLE 5.  MINIMUM AND MAXIMUM METAL DETECTION LIMITS REPORTED
          FOR TISSUE  SAMPLES  (FROM TETRA TECH 1985b)
Element
Antimony
Arsenic
Cadmium
Copper
Lead
Mercury
Nickel
Silver
Zinc
Detection Limit
(ug/g-wet weight)
Minimum Maximum
0.01 1.0
Always detected
(minimum = 0.72)
0.001 0.75
Always detected
(minimum = 0.052)
0.030 1.6
0.0004 0.09
0.019 1.0
0.001 0.27
Always detected
                                 (minimum = 1.42)
                             A-12

-------
  TABLE 6.  RECOMMENDED LIMITS OF DETECTION
  FOR WATER, SEDIMENT, AND  TISSUE MATRICES

Antimony
Arsenic
Cadmium
Copper
Iron
Lead
Mercury
Manganese
Nickel
Silver
Zinc
Water9
3
1
0.1
1
NAd
1
0.26
NA
1
0.2
1
Sedimentb
0.1
0.1
0.1
0.1
0.7
0.1
0.01
2.0
0.1
0.1
0.2
Tissue0
0.02
0.02
0.01
0.01
NA
0.03
0.01
NA
0.02
0.01
0.2
a ug/L.

D mg/kg  dry weight.

c ug/g wet weight.

d NA  =  Not applicable;  iron and manganese
used as  natural tracers for  sediments only.

e This  detection limit can be lowered  to
0.02 ug/L by using a special,  gold-amalgamation-
cold-vapor technique (Bloom and Crecelius
1983).  However, accuracy  and precision must
be  demonstrated  to the levels specified  in
the QA/QC section of the main body of this
report.
                 A-13

-------
Recommended  LOD for water  are for graphite  furnace analysis, except  for
cold vapor atomic adsorption for mercury.

SUMMARY

     U.S.  EPA priority pollutant metals  recommended for inclusion in protocol
development include antimony, arsenic,  cadmium, copper,  lead, mercury,
nickel, silver, and zinc.   U.S.  EPA priority  pollutant metals  excluded
include beryllium, chromium, selenium, and thallium.  Speciation protocols
may be useful  for selected bioaccumulation studies (i.e., arsenic and mercury),
but are not recommended for routine monitoring, or for analysis of waters
or sediments.  Iron and manganese are  recommended for inclusion  in  the
protocol  specification because of their  potential  use as geochemical tracers
and as potentially important factors in  the generation of sediment quality
values.

     The list of metals recommended for the development of sediment quality
values (SQV)  in  Puget Sound  (Tetra Tech  1985c)  includes all  13 priority
pollutant  metals.   These were presented only  as potential  candidates  for
metals of  concern.  The  purpose of the SQV  report in this  regard was  to
provide guidance for selecting contaminants of  concern rather  than to actually
select the  contaminants.  Thus, the SQV list contains some metals specifically
excluded  in this report.

     Limits of detection are recommended for the  U.S. EPA priority pollutant
metals after a consideration of attainable detection limits for each matrix
type  and  expected levels in the environment (sediments and tissues).   Based
on this comparison, maximum detection  limits  (often meaning lower cost)
are recommended that would still be  sufficient for characterizing each
matrix in a dredged materials testing program. Detection limits for speciation
analyses  (e.g., inorganic arsenic or methylmercury in tissues) are roughly
similar to those attainable for the total  metal analysis.
                                   A-14

-------
                                REFERENCES
Battelle Northwest.   1985.  Detailed  chemical and  biological  analyses of
selected sediments  from  Puget Sound.  Volume 1.  Prepared for U.S. Environmental
Protection Agency, Region  X.  Battelle Marine Research Laboratory,  Sequim, WA.
300 pp.

Bloom, N.S.,  and E.A.  Crecelius.  1983.  Determination of mercury in seawater
at sub-nanogram per  liter  levels.  Marine Chem. 14:49-59.

Crecelius, E.   22 October  1985.  Personal Communication (phone by Dr. Charles
R. Lytle).  Battelle Marine Research Laboratory,  Sequim,  WA.

Crecelius,  E.A., and C.W.  Apts.  1984.   Concentration and speciation of
arsenic in flatfish  and crabs collected from Commencement  Bay.   Tacoma-Pierce
County Health Department,  Tacoma, WA.  14 pp.

Dexter,  R.N.,  D.E.  Anderson, E.A. Quinlan, L.S. Goldstein, R.M.  Strickland,
S.P. Pavlov,  J.R. Clayton, Jr. R.M. Kocan, and M. Landolt.  1981.   A summary
of knowledge of Puget Sound related to chemical contaminants.   NOAA Technical
Memorandum OMPA-13.   National Oceanic and Atmospheric Administration, Boulder,
CO.  435 pp.

Galvin,  D.V., G.P.  Romberg, D.R. Houck, and  J.H. Lesniak.   1984.  Toxicant
pretreatment planning study.  Summary Report.   Municipality  of Metropolitan
Seattle, Seattle, WA.  202 pp.

Glaser,  J.A., D.L.  Foerst,  G.D. McKee, S.A. Quave, and W.L. Budde.   1981.
Trace analyses for  wastewaters.  Environ. Sci. Technol. 15:1426-1435.

Jones &  Stokes.  1983.   Water quality management program  for Puget Sound.
Prepared for U.S. Environmental Protection Agency, Region X.   Jones & Stokes
Associates,  Inc. Sacramento, CA.

Keith, L.J.,  W. Crummet, J. Deegan, Jr., R.A.  Libby, J.K. Taylor,  G. Weatler.
1983.  Principles of environmental analysis.  Anal. Chem. 55:2210-2218.

Konasewich,  D.E.,  P.M.  Chapman, E.  Gerencher,  G.  Vigers and N. Treloar.
1982.  Effects,  pathways, processes, and transformation of  Puget  sound
contaminants of concern.   NOAA Technical  Memorandum  OMPA-20.  National
Oceanic and Atmospheric  Administration, Boulder,  CO.  357 pp.

Malins,  D.C.,  B.B.  McCain, D.W.  Brown, A.K.  Sparks,   and  H.O. Hodgins.
1980.  Chemical contaminants and biological  abnormalities in  central and
southern Puget  Sound.  NOAA Technical Memorandum OMPA-2.  National Oceanic
and Atmospheric Administration, Boulder, CO.  295 pp.
                                    A-15

-------
Malins,  D.C.,  S.L. Chan,  B.B.  McCain, D.W.  Brown,  A.K.  Sparks,  and H.O.
Hodgins.   1981.  Puget  Sound  pollution and  its  effects on marine biota.
Progress  report  to  OMPA for the period  May 1 to September 30,  1980.  National
Marine Fisheries Service, Seattle,  WA.

Malins,  D.C.,  B.B. McCain,  D.W.  Brown, A.K.  Sparks, and H.O.  Hodgins.
1982.   Chemical  contaminants and abnormalities in fish and invertebrates
from Puget  sound.   NOAA Technical  Memorandum  OMPA-19.   National Oceanic
and Atmospheric  Administration, Boulder, CO.  168 pp.

Parson,  M.L., A. Forster, and D. Anderson.  1980.  Atlas of spectral  inter-
ferences  in  ICP  spectroscopy.   Plenum Press, New York,  NY.   644  pp.

Quinlan,  E.A.,  P.M. Chapman,  R.N.  Dexter, D.E. Konasewich,  C.C.  Ebbesmeyer,
G.A. Erickson,  B.R. Kowalski, and  T.A. Silver.  1985.   Toxic chemicals
and biological effects in Puget Sound:   states and scenarios for  the future.
Draft.   NOAA Technical Memorandum.   National Oceanic and Atmospheric Admini-
stration, Boulder,  CO. 334 pp.

Tetra  Tech.  1984.  Laboratory analytical protocol for the  Anaconda smelter
RI/FS.   Tetra Tech, Bellevue,  WA.

Tetra  Tech.  1985a.  Commencement  Bay  nearshore/tideflats  remedial  investi-
gation.  Volume  1.  Prepared  for Washington  State Department  of Ecology
and U.S.  Environmental Protection Agency, Region X.  Tetra  Tech, Inc.,
Bellevue, WA.

Tetra  Tech.   1985b.   Bioaccumulation monitoring guidance:  3.  Recommended
analytical detection limits.  Prepared  for U.S.  Environmental  Protection
Agency, Washington, DC.  Tetra Tech,  Inc., Bellevue, WA.

Tetra  Tech.   1985c.  Task 3. Evaluation of approaches for  the  development
of  sediment quality values  for Puget Sound.   Draft report prepared for
Resource Planning Associates for the U.S. Army Corps of Engineers  by  Tetra
Tech,  Bellevue,  WA.

U.S. Environmental Protection Agency.   1979.  Methods for chemical  analysis
of water and wastes.  U.S. Environmental Protection Agency Environmental
Monitoring and Support Laboratory,  Cincinnati, OH.

U.S. Environmental Protection Agency.  1980a.   Water quality criteria.
U.S. Environmental Protection Agency, Washington, DC.  Federal  Register,
Vol. 46,  No.  156, Part IV.  pp. 79318-79379.

U.S. Environmental  Protection Agency.  1980b.  Ambient water quality criterion
for mercury.  U.S.  Environmental Protection Agency, Washington,  DC.

U.S. Environmental Protection Agency. 1984.  Guidelines establishing test
procedures for the  analysis of pollutants under the clean water  act;  final
rule and interim final rule and proposed rule.  U.S. Environmental  Protection
Agency, Washington, DC.   Federal  Register,  Vol. 49, No. 209, Part VII.
pp. 199-204.

                                      A-16

-------
U.S.  Environmental  Protection Agency.   1985.  Water  quality criteria.
U.S.  Environmental  Protection Agency, Washington,  DC.   Federal Register,
Vol.  50, No. 146, Part  II.  pp. 30784-30796.
                                   A-17

-------
                        APPENDIX B






                       EXCERPTS FROM



EXHIBIT B:  REPORTING REQUIREMENTS AND DELIVERABLES



EXHIBIT D:  ANALYTICAL METHODS



EXHIBIT E:  QUALITY ASSURANCE/QUALITY CONTROL



EXHIBIT G:  CHAIN-OF-CUSTODY AND DOCUMENT CONTROL PROCEDURES



                      (U.S. EPA 1985)

-------
     CONTRACT LABORATORY PROGRAM
          STATEMENT OF WORK
                (SOW)
          INORGANIC ANALYSIS
             Multi-Media
         Mulci-Concentracion
             SOW No. 785
              July 1985
Based on:  Caucus Inorganics Prococol
                   B-l

-------
               EXHIBIT B,




REPORTING REQUIREMENTS AND DELIVERABLES
                     B-2

-------
                     REPORTING REQUIREMENTS AND DELIVERABLES

 The  contractor shall provide  reports  and other deliverables specified hereunder.
 Specific  reports  are described below.   The Contract  Reporting Schedule which
 follows the  report  descriptions gives  delivery schedule and report distribution
 Information,  Including  addresses  for  report designees.   All' reports to multiple
 addresses must be sent  out  concurrently by whatever  method necessary to meet the
 contract  required time  limits.  All reports must  also be submitted in a legible
 fora or resubmission will be  required  at no additional  cost to the Agency.

 Report Description

      A*   Weekly Progress Reports  -  Tabulation of samples  received,  date of
          receipt, and a tabulation of  problems encountered.

      B.   Sample Traffic Report  (Original Lab Copy for Return  to SMO)  - Copy  of
          SHO  Sample  Traffic Report with lab receipt  information and original
          Contractor  signature.

      C.   Sample Data Package -  Data report  package  for  analyses of  each
          sample (including  all  required QA/QC-Exhibit E) must  be complete
          before submission  and  shall Include:

          1)  Copies  of  completed  SMO Sample  Traffic  Reports with receipt of
             information completed for all samples reported in data  package.

          2}  The  cover  page for the inorganic analysis data package  (Exhibit
             B) including general comments,  Statement of Work  (SOU) Number,
             QC Report  0, sample EPA cross  reference numbers,  footnotes  used
             in the  data package, and  the  statement  on use  of  ICP  background
             and interelement corrections  for  the samples.  The SOV number
             defines  the Statement of  Work used to obtain  the  reported  values.

          3)  Tabulated  results in ug/L for aqueous samples  or  mg/kg for  solid
             samples  (identification and quantity) of specified chemical
             constituents (Exhibit C)  by the specified analyses (Exhibit 0),
             validated  and signed in original  signature  by  the Laboratory
             Manager, and reported on  Form I.*  The results for solid samples
             will be reported on a dry  weight  basis.  Percent  solids are not
             required on aqueous samples*  If the value or  the result is greacer
             than or equal to the Instrument Detection Limit (IDL), corrected for
             dilutions,  as determined in Exhibit E, report  the value and Indicate
             the analytical method used  for  the metals*  Use P_ for ICP, A_ for
             flame AA and J_ for furnace AA.  All dilutions not* required by the
             contract and Tffectlng the  IDL, must be noted  on  an element by ele-
             ment basis  on Form I.  If  the value is less than  the Contract Re-
             quired Detection Limits (CRDL)  in Exhibit C, put  the value  in
             brackets, e.g.   lib).  If  the element was analyzed for but not de-
             tected, report the instrument detection limit  value with a  "IT
*In che event the Laboratory Manager cannot validate all data reported for
 each sample, he/she will provide a detailed description of the problems
 associated with the sample.
                                        B-3

-------
    (e.g.. LOU).  Use an "E" as the footnote Co indicate an estimated
    value or value noc reported due to the presence of interference,
    and an explanatory note oust be included on the cover page.  If
    the duplicate sample analysis is not within the control limits,
    flag it with an asterisk (*}.  Use "S" as a footnote to indicate
    a value determined by Method of Standard Additions (MSA).  If the
    correlation coefficient (r) for method of standard additions is
    less than 0.995, flag the value with a '+' sign.  If duplicate
    injection precision criteria specified for Furnace AA analysis in
    Exhibit E cannot be net, flag the data with an "M".  If the spike
    sample recovery is not within control limits, flag the data with
    the letter N.  Report results to two significant figures for values
    from 0 to 100 and three significant figures for results greater
    than 100, with the exception of Mercury (see Mercury Methods -
    Exhibit D).  For rounding rules, follow the EPA Handbook of
    Analytical Quality Control in Water and Wastewater Laboratories
    (EPA-600/4-79-019).

    Under the comments section on Form I, provide a brief physical
    description of the sample using the following guidelines:

    A.   Water samples - Coloration and clarity

    B.   Solid samples - Coloration, texture and artifacts

    Recommended Descriptive Terms

         Coloration;  Red, blue, yellow,  green, orange, violet, white,
              colorless, brown, grey, black

         Clarity:  Clear, cloudy, opaque  .

         Texture;  Fine (powdery), Medium (sand),  Coarse (large crystals
              or rocks)

    Also note any significant changes that occur during sample prepara-
    tion (i.e. coloration shifts, emulsion formation).

4)  Analytical results for samples and spikes, duplicates, standards,
    ICP Interference Check Samples, reagent blanks, laboratory control
    samples,  Instrument detection limits  and holding times on-QA Forms
    II, III,  IV, V, VI, VII, VIII, IX and X.   Multiple forms require
    identification (i.e.  Form IIA, Form IIB etc.)  Summarize each full
    method of standard addition performed on Form VIII.

5)  Legible photocopy of raw data (sequential measurement readout
    record) clearly labeled with sufficient information to unequivocally
    Identify:

    a)  calibration standards (including  prep date).

    b)  calibration blanks and preparation blanks.

                               B-4

-------
             c)  initial and continuing calibration verification standards and
                 Interference check samples.

             d)  diluted and undiluted samples (by EPA number) and all weights,
                 dilutions and volumes used to obtain the reported values.   If
                 the volumes, weights and dilutions are consistent for all sam-
                 ples in a given Case, a general statement outlining these
                 parameters is sufficient.

             e)  duplicates.

             f)  spikes (indicating standard solutions used, final spike concen-
                 trations, volumes involved)*

             g)  any instrument adjustments, data corrections or other apparent
                 anomalies on the measurement record including all data voided
                 or data not used to obtain reported values*

             h)  all information for furnace analysis contained in Fora VIII
                 (Exhibit E), clearly identified on the raw data including
                 sample #, initial single spike data, Z recovery, full MSA data,
                 MSA correlation coefficient, slope and intercept of linear fit,
                 and final sample concentration (standard addition concentra-
                 tion) .

         6)  Copies of digestion logs for the 1C?, flameless AA and Eg prepar-
             ations and of the distillation log for cyanide.  These logs must
             Include:  1) date,  2) sample weights and volumes, 3) sufficient
             information to unequivocably identify which QC samples (i.e. LCS,
             preparation blank) correspond to each batch digested, 4) comments
             describing any significant sample changes or reactions which occur
             during preparation, 5) indication of pH <2 or >12, as applicable.

         7)  The order of raw data in the data package shall be:  1) ICF, 2)
             Flame AA, 3) Furnace AA, 4) Hg, 5) CN, 6) Digestion and Distilla-
             tion logs, 7) Percent Solids.  All raw data must include intensities
             (1C?) and absorbances (AA) unless instrument direct readout is in
             concentration units.

         8)  For each reported value, the data package must contain all raw data
             from the instrument used to obtain that value and the QA/QC values
             reported.  All instruments must provide a hard copy of the instru-
             ment readout (i.e.  stripcharts, printer tapes, etc.).  A photocopy
             of the instrument readout must be included in the raw data package.

    Information shall include a key to abbreviations, with response units
    stated, and a cross reference to EPA sample numbers.

D.  Results of Intercoaparison/Perfomance Evaluaclon(PE) Sample Analyses

    Tabulation of analytical results for Interconparlson/PE Sample analyses
    include all requirements specified in C above.
                                        B-5

-------
E.  Complete Case File Purge (formerly termed the "Document Control and Chain-of-
    Custody Package")

    The Complete Case File Purge package must include all laboratory records
    received or generated for a specific case that have not been previously
    submitted to EPA as deliverable.  These items include but are not limited
    to:  sample tags, custody records, sample tracking records, analysts logbook
    pages, bench1 sheets, instrument readout records, computer printouts, raw
    data summaries, instrument logbook pages (including instrument conditions),
    correspondence, and the document inventory.

    Shipment of each Complete Case File Purge package by first class mail,
    overnight carrier, priority mail or equivalent is acceptable.  Custody
    seals, which are provided by EPA, must be placed on shipping containers and
    a document inventory and transmittal letter included.  The laboratory is
    not required to maintain any documents for a case after shipment; however,
    it is recommended that you maintain a copy of the document inventory and
    transmittal letter.

F.  Quarterly Verification of Instrument Parameters

    The Contractor muse perform and report quarterly verification of instrument
    detection limits by methods specified in Exhibit E and report type and
    model 0 for each instrument used on this contract.  For the 1CP instrument-
    ation and methods, the Contractor must also report quarterly: linearity
    range verification, interelement correction factors, wavelengths used and
    integration times.  This information is reported using Forms XI, XII and
    XI11.  Submissions of Quarterly Verification of Instrument Parameters oust
    include the raw data used to determine those values reported.

G.  Results of Solid Laboratory Control Samples (LCS)

    Tabulation of analytical results and QC for the solid LCS sample analysis,
    as specified in Exhibit E.
                                     B-6

-------
                          CONTRACT REPORTING SCHEDULE
Report
A. Weekly Progress
Report
B. Sample Traffic
Re pore
C. Sample Data
Package
D* Results of Incer-
comparison Study/
FE Sample Analysis
E. Compilation of
Complete Case File
Purge Package
F. Complete Case
File Purge
Package
C. Quarterly
Verification
of Instrument
Parameters
H. Results of Solid
LCS Analysis
Mo.
1 Copies
1
1
3
1
I
1
2
2
Delivery
Schedule*
Weekly
7 days f roo
receipt of
sample
35 days from
receipt of
sample
35 days from
receipt of
sample
7 days after
data submission
180 days after
data submission
or 7 days from
request by PO
or SMO
Quarterly:
15th day of
January, April,
July, October
Monthly: 15th
of every month
SMO( 1 )
X
X
X



X
f*
A
ENSL/
1 LV(2>


X
X


X
X
1 Region/
J Clienc(3)


X





ICEAT
UEIC(4)





X


•Schedule designated in calendar days.
                                       B-7

-------
REPORT DISTRIBUTION ADDRESSEES;

(1)  USEPA
     Contract Laboratory Program
     Sample Management Office (SMO)
     P.O. Box 318
     Alexandria, VA 22313
                                      (2)  USEPA
                                           Environmental Monitoring
                                            Systems Laboratory (EI1SL/LV)
                                           P.O. Box 15027
                                           Us Vegas. NV 8911*
                                           AITN:  Data Audit Staff
     For overnight delivery service,
     use street address:

     300 H. Lee Street
     Alexandria, VA  22314
(3)  USEPA REGIONS:
                                           For overnight delivery service,
                                           use street address:

                                           944 E. Harmon, Exec. Ccr. Ha.  226
                                           Us Vegas, NV  49109
                                           ATTN:  Data Audit Scaff
(4)
The CLP Sample Management Office will provide the contractor with the list
of addresses for distribution to the ten EPA Regions.  SMO will provide
the Contractor with updated Regional address/name lists as necessary
throughout the period of. the contract*

USEPA
HE1C Contract Evidence Audit Team
12600 West Coifax Ave., Suite C310
La lie wood, CO 802 IS
                                     B-3

-------
U.S. EPA Cone race Laboratory Program
Sample Management Office
P.O. Box 818 - Alexandria, VA  22313
703/557-2490  FTS: 8-557-2490                                 Date 	

                                    COVER PACE
                         INORGANIC ANALYSES DATA PACKAGE

Ub Name 	                  Case No. 	
SOW No.  	                  Q«C. Report No. 	

Ub Receipt Date 	
                                  Sample Numbers
EPA No.               Lab ID No.              EPA No.             Lab ID No.
Comments:
ICP interelement and background corrections applied?  Yes 	  No 	.
Lf yes, corrections applied before _____ or after	generation of raw data.

Footnotes:
MR   -  Net required by contract at this time
Form I:
Value - If the result is a value greater than or equal to the instrument
        detecion limit but less than the contract-required detection limit,
        report the value in brackets (i.e.,  [10]).  Indicate the analytical
        method used-with V (for ICP),  A (for Flame AA) or. F (for Furnace AA).
U     - Indicates element was analyzed for but not detected.  Report with che
        instrument detection limit value (e.g.,  10U).
E     - Indicates a value estimated or not reported due to the presence of
        interference.  Explanatory note included on cover page.
s     - Indicates value determined by Method of Standard Addition.
U     - Indicate? spike sample  recovery is not within control limits.
*     - Indicaces duplicate analysis Is not within control limits.
+     - Indicates the correlation coefficient for method of standard addition is
        less than 0.995
M     - Indicates duplicate  Injection results exceeded control limits.

Indicate  method used: P for  ICP; A for Flame AA and F for Furnace.
                                       B-9

-------
                                      Form I
U.S. EPA Contract Laboratory Program
Sample Management Office
P.O. Box 818 - Alexandria, VA  22313
703/557-2490  FTS: 8-557-2490
                                               EPA Sample No.
                                                          Date
LAB NAME 	

SOW NO.  	

LAB SAMPLE ID. NO.
              INORGANIC ANALYSIS DATA SHEET

                                     CASE NO.
                                     QC REPORT NO.
                         Elements Identified and Measured
Concentration: Low
Matrix: Water

Soil
Medium
Sludge Other
ug/L or mg/kg dry weight (Circle One)
1* Aluminum 13. Magnesium
2. Antimony
3. Arsenic
4. Barium
5. Beryllium
6. Cadmium
7. Calcium
8. Chromium
9. Cobalt
10. Copper
11. Iron
12. Lead
14.
15.
16.
17.
18.
19.
20.
21.
22.
23.
Manganese
Mercury
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc
Precent Solids (Z)
Cyanide
Footnotes:
For reporting results to EPA, standard result qualifiers are used
as defined on Cover Page.  Additional flags or footnotes explaining
results are encouraged.  Definition of such flags oust be explicit.
and contained on Cover Page, however.
Comments:
                                               Lab  Manager
                                         B-10

-------
                                      Fora II
                             Q. C. Report No.
LAB NAME
DATE
 INITIAL AND CONTINUING CALIBRATION VERIFICATION3
	               CASE NO. 	
                             SOW NO.  	
                             UNITS
Compound Initial Calib. 1 Continuing Calibration2
Metals:



5. Beryllium



Cobalt


12. Lead



16. ' Nickel


19. Silver
20. Sodium
21. Thallium
22. Vanadium
23. Zinc
Other:

inide
True Value














.




m






Found






•















•



ZR









•













•


True Value













•












Found


























ZR


























Found















•










ZR


























Method4


























 1  Initial  Calibration  Source
                         2 Continuing Calibration Source_
 3  Control  Limits:  Mercury  and  Tin  80-120;      Other Metals 90-110;  Cyanide 85-115
 4  Indicate Analytical Method Used:    P - ICP;  A - Flame AA;  F - Furnace AA

-------
LAB NAME
DATE
                                      Fora  III
                              Q. C. Report  No.
                                     BLANKS
CASE NO,
UNITS
                                 Matrix
Compound
Metals:
1 • Aluminum
2. Antimony
3, Arsenic
A. Barium
5. Beryllium
6. Cadmium
7. Calcium
8. Chromium
9. Cobalt
10* Copper
11. Iron
12. Lead
13. Magnesium
14. Manganese
15. Mercury
16. Nickel
17. Potassium
18. Selenium
19. Silver
20 • Sodium
21. Thallium
22. Vanadium
23. Zinc
Other:
'
Cyanide
Initial
Calibration
Blank Value









*

•














Continuing Calibrate
Blank Value
1























i


2 3




















































m
4









•








•







Preparation Blank
Matrix: Matrix:
1 2










•


•












                                          B-12

-------
LAB  NAME
DATE
                                       Fora IV
                              Q.  C.  Report No.
ZCP INTERFERENCE CHECK SAMPLE
                        CASE NO.
                                                    Check Sample I.D.
                        Cheek Sample Source
                        Units
Compound
Metals:
1. Aluminum
2. Antimony
3. Arsenic
A. Barium
5. Beryllium
6. Cadmium
7. Calcium
8. Chromium
. Cobalt
10* Copper
11* Iron
12. Lead
13. Magnesium
14. Manganese
15. Mercury
16. Nickel
17. Potassium
18. Selenium
19. Silver
20. Sodium
21. Thallium
22. Vanadium
23. Zinc
Other:

Control Limit si
Mean

























Std. Dev.

























True2


•





.
















Initial
Observed

























ZR

























Final
Observed

























ZR

























*  Mean value based on n  •
   True value of EPA ICP Interference Check Sample or contractor standard.
                                         B-13

-------
                                        Fora  V

                               Q.  C.  Report No.
                                SPIKE SAMPLE  RECOVER*
LAB NAME

DATE
          CASE NO. 	
          EPA Sample No. 	
          Lab Sample ID No.
          Units
                                Matrix
Compound
Metals:
1 . 'Aluminum
2t Antimony
3* Arsenic
4. Barium • •
5* Beryllium
6 . Cadai un
7 . Calcium
8. Chromium
9. Cobalt
10. Copper

12. Lead
13. Magnesium
14. Manganese
15. Mercury
16. Nickel
17 i Potassium
18. Selenium
19. Silver
20. Sodium


23. Zinc
Other:


Control Limit
ZR
75-125
»
•
_ __ m
•
•
m
•
-
•
•
H
N
-
•
•
«•
*•
-
-
-
W
-


M
Spiked Samole
Result (SSR)



















•






Sample
Result (SR)


























Spiked
Added (SA)
•
















•


•





ZRl


























 1  ZR  "  USSR -  SR)/SA]  x 100

 "N"-  cue  of  control
"MR*-  Not  required
 Comments:
B-L4

-------
LAB NAME

DATE
                                      Fora VI
                                 C. Report No.
                                     DUPLICATES
         CASE NO. 	
         EPA Sample No. 	
         Lab Sample ID No.
         Units    	
                                Matrix
Compound
Metals:
1 . Aluminum
2. Antimony
3. Arsenic
4. Barium
5. Beryllium
6. Cadmium
7. Calcium
8. Chromium
9. Cobalt
J. Copper
11. Iron
12. Lead
13. Magnesium
14. Manganese
15. Mercury
16. Nickel
17. Potassium
18. Selenium
1*». Silver '
20. Sodium
21. Thallium
22. Vanadium
23. Zinc
Other:

ranide
Control Limit*
•





.














•

•


Sample(S)





















B




Dupllcate(D)






















.



RPD2


























* Out of Concrol

1 To be added at a later date.
2 RPD • [|S - D|/((S +  D)/2)]  x 100
NC - Non calculable RPD due to value(s) less  than CRDL
                                          B-15

-------
                                      Form VII
                             Q.C. Report No.
LAB NAME
INSTRUMENT DETECTION LIMITS AND
   LABORATORY CONTROL SAMPLE
          CASE NO.      	
DATE
                                                             LCS NO.
Compound
Metals:
1. Aluminum
2. Antimony
3* Arsenic
4. Barium
5. Beryllium
b« Cadmium
7. Calcium
8. Chromium
9. Cobalt
10. Copper
11. Iron
12. Lead
13. Magnesium
14. ilanganese
15. Mercury
16. Nickel
17. Pocassiua
18. Selenium
19. Silver
20. Sodium
21. ThalliuD
22. Vanadium
23. Zinc
Other:

Cyanide
Required Dececcion
Limits (CRDL)-ug/!

200
60
10
200
5
5
sooo
10
so
25
100
5
5000
15
0.2
40
5000
5
10
5000
10
50
20


10
Instrument Detection
Limits (IDL)-ug/!
ICP/AA Furnace
ID# 	 ID* 	













•

-









NR

























f.'R
Lab Control Sample
ug/L mg/kg
(circle one)
True Found ZR










































•



































NR - Noc  required
                                       B-16

-------
                                     Fora VIII
                             Q.C. Report No. __
                             STANDARD ADDITION RESULTS
LAB NAME
DATE
CASE NO.
UNITS
EPA
Sample 9






















Element




•




—


•



—




—
0 ADD
ABS.






















1 ADD
CON.






















ABS1






















2 ADD
CON.






















ABS.1






















3 ADD
CON.






















ABS.1






















FINAL
CON. 2






















r*








•













 1  CON  is  the  concentration added.. ABS.  is the instrument readout in absorbance or
   concentration.
 2  Concentration as  determined by MSA
 *"r" is  the correlation  coefficient.
 +  - correlation coefficient is outside  of control window of 0.995.
                                          B-17

-------
                                     Form IX

                             Q. C. Report No.

                                ZCP SERIAL DILUTIONS
LAB NAME
CASE NO.
                                                    EPA Sample No.
DATE
Lab Sample ID No.

Units
                               Matrix
Compound
Metals:
1 . Aluminum
Initial Sample
Concentration I)

Serial Dilution1
Result (S)

Z Difference2

2. Antimony
3* Arsenic
4. Barium
5. Beryllium
6» Cad^yp
7. Calcium
8 • Chromium
9. Cobalt
10* Copoer
11. Iron
12. Lead
13. Magnesium
14. Manganese
15. Nickel
16. Potassium
17. Selenium
18. Silver
19. Sodium
20. Thallium
21. Vanadium
22. Zinc
Other:




|

































































1 Diluted sample concentration corrected for 1:4 dilution (see Exhibit D)
2 Percent Difference •   I1 " sl   x  100
                           I
NR - Not Required, initial sample concentration less than 10 times IDL
NA - Not Applicable, analyte not determined by ICP
                                           B-18

-------
                                      Form X
                             QC Report No. 	
                                   HOLDING TIMES
LAB NAME
DATE
CASE NO.
EPA
Sample No*



























Matrix



























Date
Received





•





















Mercury
Prep Date



























Mercury
Holding Time
(Days)



























CN Prep
Date








.


















CN
Holding Time
(Days)














.












                                        B-19

-------
                               Fora XI (Quarterly)
                           INSTRUMENT DETECTION LIMITS
LAB NAME
DATE
ICP/Flame AA (Circle One) Model Number
Furnace AA Number
Element
1 • Aluminum
2. Antimony
3. Arsenic
4. Barium
5. Beryllium
6. Cadmium
7. Calcium
8. Chromium
9. Cobalt
10. Copper
11. Iron
12. Lead
Wavelength
(nm) J












CXDL
(ug/L)
200
60
10
200
5
5
5000
10
50
25
10U
5
IDL
(ug/L)

























Element
13. Magnesium
14. Manganese
15. Mercury
16. Nickel
17. Potassium
18. Selenium
19. Silver
20. Sodium
21. Thallium
22. Vanadium
23. Zinc
Wavelength

-------
                                   Fora XIZ (Quarterly)
                         ICP InterelemcQt Correction Factors
 LABORATORY,
 DATE
ZCP Model Number

Analyte
1. Antimony





7. Cobalt
8. Copcer
9. Lead
10* Manganese

12. Nickel


IS. Silver


IS. Vanadium
19. Zinc
Analyte
Wavelength
(na)











•







Intereleaent Correction Factors
for
Al



















Ca
•


















. Fe



















Mg



































































































COMMENTS:
                                   B-21
                                             Manager

-------
                              Fora XII (Quarterly)  (cont'd)
                         ICP Ineerelemenc Correction Factors
 LABORATORY^

 DATE
ICP Model Number

Analyte
1. Antimony


4. Beryllium
5. Cadmium
6. Chromium
7. Cobalt
8. Copper
9. Lead


12. Nickel


15. Silver



19. Zinc
Analyte
Wavelength
(nm)





•


•










Interelement Correction Factors .
for


































































•









































•



















































COMMENTS:
                                         Lab Manager
                                         B-22

-------
                                 Fora XIII  (Quarterly)
                                   ICP Linear Ranges
LAB NAME

    DATE
ICP Model Number
Analyte
1. Aluminum
2. Antimony
3. Arsenic
4. Barium
5. Beryllium
6. Cadmium
7. Calcium
8. Chromium
9. Cobalt
10. Copoer
11. Iron
12. Lead
Integration
Time
(Seconds)












Concen-
tration
(us/L)












Aaalyte
13. Magnesium
14. Manganese
15. Mercury
16. Nickel
17. Potassium
IS. Selenium
19. Silver
20. Sodium
21. Thallium
22. Vanadium
23. Zinc
Integration
Time
(Seconds)











Concen-
tration
(ui/L)












Footnotes:        •  Indicate elements not analyzed by ICP with the notation "NA",
COMMENTS:
                                           Lab Manager
                                      B-23

-------
    EXHIBIT D




ANALYTICAL METHODS
       B-24

-------
                               Analytical Methods

Any analytical method specified in Exhibit D may be utilized as long as the
documented instrument or method detection limits meet the contract required
detection levels (Exhibit C).  Analytical methods with higher detection limits
may be used only if the.sample concentration exceeds twice the documented
detection limit of the instrument or method.  When an analyte concentration
exceeds the calibrated or linear range, reanalysis of the prepared sample is
required after appropriate dilution.  All samples must initially be run un-
diluted (i.e. final product of sample preparation procedures).  Both diluted
and undiluted sample measurements must be contained in the raw data.

Labware must be acid cleaned according to EPA's manual "Methods for Chemical
Analysis of Water and Wastes'  (2) or as equivalent procedure.  Samples must be
opened and digested in a hood.  Stock solutions for standards may be purchased
or made up as specified in the analytical methods.  Deionized water (Type II)
is equivalent to distilled water.  All sample dilutions should be made with
acidified water to maintain constant add strength.  Background corrections are
required for flame AA measurements below 350 na and all furnace AA measurements.
Before water sample preparation is initiated, check the pH of all water samples
and note in the sample preparation log if it is less than 2 for metals and
greater than 12 for cyanide.

NOTE:   Reference numbers appear in parentheses throughout this section.
        See the last page of  this exhibit for reference list.

1.  Furnace AA/IC?/Flame AA - Water

    Furnace Atomic Absorption

    A.  Sample Preparation  (Furnace Digestion Procedure for Waters)

        Shake  sample and  transfer  100 mL of well-mixed  sample  to  a  250 mL
        beaker, add  1 mL  of  (1+1) HN03  and  2 mL 30Z H202.  Cover with watch
        glass  or  similar  cover, heat for 2  hours at 95*C or until the volume
        is  reduced  to- between -25  and 50 mL  (make  certain samples  do not boil).
        Cool  sample  and filter (see Mote  1) to  remove insoluble material and
         bring back  to  100 mL with  deionized distilled water.   The sample is
        now ready for  analysis.

        Concentrations  so determined shall  be  reported  as  "total".

         If  Sb is  to be determined by  furnace,  use the  dlgestates  prepared  for
         ICP/AA analysis.

     B.  Analytical Procedures

         See Attachment 1 (3).i

         Each furnace analysis requires at least 2 burns,  except when performing
         full method of standard additions (MSA).
                                       B-25

-------
   Inductively Coupled  Plasma  (ICP)  and  Plane  Atomic Absorption (AA)

   A.   Sample Preparation (ICP/Flame AA  Digestion Procedure for Water)

        Shake sample and transfer 100 mL  of:a veil mixed sample to a beaker.
        Add 2 mL  of  (1+1) HN03  and 10'mL  of (1+1)  HC1 to the sample.  Cover with
        watch glass  or similar  cover  and  heat on a steam bath or hot plate until
        the volume has been reduced to between  25 and 50 mL making certain the
        sample does  not  boil.   After  this treatment,  cool sample and filter to
        remove insoluble material that could clog the nebulizer.  (See Note 1.)
        Adjust the volume to 100 mL with  deionized distilled water.  The sample
        is now ready for analysis.

        Concentrations so determined  shall be reported as "total."

        NOTE 1:   In  place of filtering, the sample after dilution and mixing
                 may be centrifuged or allowed  to settle by gravity overnight
                 to  remove insoluble  material.

    B.   Analytical Procedures

        ICP      - See Attachment 3.
        Flame AA - See Attachment 4.

2.  Furnace AA/ICP/Flame AA - Sediment, Sludges and Soils

    A.   Sample Preparation - Acid Digestion of Sediments, Sludges and Soils

        See Attachment 1.

    B.   Analytical Procedures

        Furnace  AA - See Attachment 2.
           Each furnace analysis requires at least 2 burns, except for full MSA.
        ICP        - See Attachment 3.
        Flame AA   - See Attachment A.

3.  Mercury  in Water
    See Attachment 5  (manual) and 5A  (automated).

4.  Mercury  in Sediment
    See Attachment 6.

5.  Cyanide  in Water
    See Attachment 7.

6.  Cyanide  in Sediment
    See Attachment 8.

7.  Determination of  Percent Solids
    See Attachment  9.


                                      B-26

-------
                                  ATTACHMENT 1

               SAMPLE PREPARATION OF SEDIMENTS, SLUDGES AND SOILS

1.  Scope and Application

    i.i  This method is an acid digestion procedure used Co prepare sediments,
         sludges, and soil samples for analysis by flame or furnace atomic
         absorption spectroscopy (AAS) or by inductively coupled argon plasma
         spectroscopy (1C?).  Samples prepared by this method may be analyzed
         by AAS or ICP for the following metals:

                 Aluminum        Chromium        Potassium
                 Antimony        Cobalt          Selenium
                 Arsenic         Copper          Silver
                 Barium          Iron            Sodium •
                 Beryllium       Lead            Thallium
                 Cadmium         Magnesium       Vanadium
                 Calcium         Manganese       Zinc
                                 Nickel

2.  Summary of Method

    NOTE:  A separate digestion procedure is required for furnace AA and ICP
           analysis.

    2.1  A representative 1 g (wet weight) sample is digested in nitric acid and
         hydrogen peroxide.  The dlgestate is then refluxed with either nitric
         acid or hydrochloric acid.  Hydrochloric acid is used as the final
         reflux acid for the furnace AA analysis of Sb, the flame AA or ICP
         analysis of Al, Sb, Ba, Be, Ca, Cd, Cr, Co, Cu, Fe, Pb, Mg, Mn,
         Nit K, Ag, Na, Tl, V and Zn.  Nitric acid is employed as the final
         reflux add for the furnace AA analysis of Aa, Be, Cd, Cr, Co, Cu, Fe,
         Pb, Mn, Ni, Se, Ag, Tl, V, and Zn.  A separate sample shall be dried
         for a total solids determination (Exhibit D, Attachment 9).

3.  Apparatus and Materials

    3.1  250 ml beaker or other appropriate vessel.

    3.2  Watch glasses

    3.3  Thermometer that covers range of 0° to 200*C

    3.4  Whatman No. 42  filter paper  or equivalent

4.  Reagents

    4.1  ASTM Type  II water  (ASTM D1193):  Hater must be monitored.

    4.2  Concentrated  Nitric Acid  (sp. gr.  1.41)


                                     B-27

-------
     4.3   Concentrated Hydrochloric Acid (sp.  gr.  1.19)

     4.4   Hydrogen Peroxide (30%)

5.   Sample Preservation, and Handling

     5.1   Non-aqueous samples muse be. refrigerated upon receipt uncil analysis.

6.   Procedure

     6.1   Mix Che sample thoroughly to achieve homogeniety.  For each digestion
           procedure, weigh  (to the nearest O.Olgns) a 1.0 to 1.5 gm portion of
           sample and transfer to a beaker.
     6.2   Add  10 ml of  1:1 nitric acid (HNC^), mix the slurry, and cover with
           a watch glass.  Heat the sample to 95 °C and reflux for 10 minutes
           without boiling,  Allow the sample to cool, add 5 ml of concentrated
           HN03, replace the watch glass, and reflux for 30 minutes.  Do not
           allow the volume to be reduced to less than 5 ml while maintaining
           a covering of solution over the bottom of the beaker.

     6.3   After the second reflux step has been completed and the sample has
           cooled, add  2 ml of Type II water and 3 ml of 30Z hydrogen peroxide
           (H202>*  Return the beaker to the hot plate for warming to start the
           peroxide reaction.  Care must be taken to ensure that losses do not
           occur due to excessively vigorous effervescence.  Heat until effer-
           vescence subsides, and cool the beaker*

     6.4   Continue to  add 30Z H202 in 1 ml aliquot s with warming until the
           effervescence is minimal or until the general sample appearance is
           unchanged.   (NOTE:  Do not add more  than a total of  10 ml  302 H202. )

     6.5    If  the  sample is  being prepared for  the furnace AA  analysis of  Sb,
            the flame AA or 1CP analysis of Al,  Sb. Ba, Be, Ca,  Cd,  Cr, Co, Cu,
            Fe, Pb,  Mg,  Hn, Hi, K, Ag,  Ha, Tl,  V, and  Zn, add 5 ml of  1:1 HC1
           and 10  ml of Type  II  water, return  the covered beaker  to the hot
            plate,  and  heat for an additional  10 minutes.  After cooling, filter
            through Whatman No.  42 filter  paper (or equivalent) and  dilute  to
            100 ml  with Type II  water (or  centrifuge  the  sample -  see Note  1).
            The diluted sample has an approximate acid concentration of  2.5*
            (v/v) HC1 and 5Z (v/v) HN03.   Dilute the  dlgestate  1:1  (200 ml  final
            volume) with the delonized water.   The  sample  is  now ready for
            analysis.

      6.6   L£  the  sample  is being prepared for the  furnace  analysis of  As, Be,
            Cd, Cr, Co,  Cu, Fe,  Pb,  Hn. Ni,  Se, Ag,  Tl,  V,  and  2n,  continue heat
            ing the acid-peroxide digestate until  the volume has been reduced  to
            approximately  2 ml, add 10 ml of  Type  II water,  and warm the  mixture
            After cooling, filter through Whatman  No.  42 filter paper (or equi-
            valent - see Note 1) and dilute to  100 ml with Type 11 water (or
            centrifuge  the sample).   The diluted digestate solution contains
                                     B-28

-------
          approximately 22 (v/v) HN03.   Diluce  the  digestace  1:1  (200 ml final
          volume) with deionized water.   For analysis,  withdraw aliquocs of
          appropriate volume,  and add any required  reagent  or matrix modifier.
          The sample is now ready for analysis.

7.    Calculations

     7.1  A separate determination of percent solids  oust  be  performed
          (Exhibit D, Attachment 9).

     7.2  The concentrations determined  in the  digest are  to  be reported
          on the basis of the dry weight of the sample.

     Concentration (dry wt.) (mg/kg)  .  C x V
                                        wHTs

      where   C •  Concentration  (mg/L)
              V •  Final volume in liters after sample preparation
              W »  Weight in kg of wet sample
              S -  Z Solids/100

     R£F:  Modification of Method 3050 from. SW-846, Test Methods  for Evaluating
     Solid Waste, EPA Office of Solid Waste and Emergency  Response, July  1982.


8.    Bibliography

     1.  Modification (by committee)  of  Method 3050,  SW-846,  2nd  ed., Tesc
         Methods for Evaluating Solid Waste. EPA Office of  Solid  Waste  and
         Emergency Response, July 1982.
                                    B-29

-------
                                  ATTACHMENT 2

                                    ANTIMONY

           Method 204.2 CLP-M (Atomic Absorption, furnace technique)

Optimum Concentration Range:  20-300 ug/1
Approximate Detection Limit:  3 ug/1

Preparation of Standard Solution

     1.  Stock solution:  Carefully weigh 2.7426 g of antimony potassium
         tartrate (analytical reagent grade) and dissolve in delonlzed
         distilled water.  Dilute to 1 liter with deionized water.
         1 mL - 1 mg Sb (1000 og/L).

     2.  Prepare dilutions of the stock solution to be used as calibration
         standards at the time of analysis.  These solutions are also to be
         used for "standard additions".

     3.  The calibration standards must be prepared using the same type of
         acid and at the same concentration as will result in the sample to
         be analyzed after sample preparation.

Instrument Parameters (General)

     1.  Drying Time and Temp:  30.sec-125'C.
     2.  Ashing Time and Temp:  30 sec-800°C.
     3.  Atomizing Time and Temp:  10 sec-2700'C.
     4.  Purge Gas Atmosphere:  Argon
     5.  Wavelength:  217.6 no
     6.  Other operating parameters should be set as specified by the
         particular'instrument manufacturer.
Notes
      1.  The above  concentration values and instrument  conditions  are  for  a
         Perkin-Elmer HGA-2100, based  on  the use  of  a 20  ul  Injection,  contin-
         uous  flow  purge gas and non-pyrolycic graphite and  are  to be  used as
         guidelines only.   Smaller  size furnace devices or those employing
         faster  rates of atomlzatlon can  be operated using lower atomizaclon
         temperatures for  shorter time periods than  the above  recommended
        .-settings.

      2.  The use of background  correction is required.

      3.  Nitrogen may also be used  as  the purge gas.

      4.  If  chloride concentration  presents a matrix problem or  causes a loss
         previous to atomlzation, add  an  excess 5 mg of ammonium nitrace Co
         the furnace and ash using  a ramp accessory  or  with  incremental seeps
         until the  recommended  ashing  temperature is reached.
                                       B-30

-------
5.  For every sample anal/zed, verification is necessary Co determine
    that method of standard addition is not required (see Exhibit E).

6.  If method of standard addition is required follow the procedure
    given in Exhibit E.
                                  B-31

-------
                                    ARSENIC

           Method 200.2 CLF-M (Atomic Absorption, furnace  technique)

Optimum Concentration Range:  5-100 ug/1
Approximate Detection Limit:  1 ug/1

Preparation of Standard Solution
     1.  Stock solution:  Dissolve 1.32U g of arsenic trioxide, As203
         (analytical reagent grade) in  100 ml of delonlzed distilled water
         containing 4 g NaOH.  Acidify  the solution with 20 ml cone. UNO3
         and dilute to 1 liter.  1 ml • 1 mg As (1000 mg/1).
     2.  Nickel Nitrate Solution, 52:   Dissolve 24.780 g of ACS reagent grade
         Ni(N03)2*6H20 in delonlzed distilled water and make up to  100 ml.
     3.  Nickel Nitrate Solution, 12:   Dilute 20 ml of the 52 nickel nitrate
         to 100 ml with deionized distilled water.
     4.  Working Arsenic Solution:  Prepare dilutions of the stock  solution to
         be used as calibration standards at the time of analysis.  Withdraw
         appropriate allquots of the stock solution, add 1 ml of cone. HN03,
         2 ml of 302 &202 and 2 ml of the 52 nickel nitrate solution.  Dilute
         to 100 ml with deionized distilled water.

Sample Preparation

     1.  Add 100 ul of the  52 nickel nitrate solution to 5 ml of the digested
         sample.  The sample*is now ready for injection into the furnace.

Instrument Parameters (General)
     1.  Drying Time and Temp:  30 sec-125°C.
     2.  Ashing Time and Temp:  30 sec-1100°C.
     3.  Atomizing Time and Temp:  10 sec-2700aC.
     4.  Purge Gas Atmosphere:  Argon
     5.  Wavelength:  193.7 nm
     6.  Other operating parameters should be set as specified by the
         particular Instrument manufacturer.
Notes
      1.  The  above  concentration  values  and  Instrument  conditions  are  for  a
         Perkln-Elmer HGA-2100, based  on the use of  a 20 ul injection,  purge
         gas  interrupt and  non-pyrolytic graphite.   Smaller size furnace
         devices  or those employing faster rates of  atomlzation can be  operated
         using lower atomlzation  temperatures for shorter time  periods  than  the
         above recommended  settings.

      2.  The  use  of background correction is required.

      3.  For  every  sample analyzed, verification is  necessary to determine
         that method of standard  addition is not required (see  Exhibit  E).

      4.  If method  of standard addition  is required,  follow the procedure
         given in Exhibit E).

      5.  The  use  of the Electrodeless  Discharge  Lamps (EDL) for the light
         source is  recommended.       B-32

-------
                                    CADMIUM

           Method 213.2 CLP-M (Atomic Absorption, furnace technique)

Optimum Concentration Range:  0.5-10 ug/1
Approximate Detection Limit:  0.1 ug/1

Preparation of Standard Solution

     1.  Stock solution:  Carefully weigh 2.282g of cadmium sulfate, 3 Cd
         8 B20 (analytical reagent grade) and dissolve in delonized distilled
         water.  Make up to 1 liter with delonized distilled water.  1 mL »
         1 mg Cd (1000 mg/L).

     2.  Ammonium Phosphate solution (40Z):  Dissolve 40 grams of ammonium
         phosphate, (NH^ZHPO^ (analytical reagent grade) in delonized distilled
         water and dilute to 100 ml.

     3.  Prepare dilutions of stock cadmium solution to be used as calibration
         standards at the time of analysis.  To each 100 ml of standard and
         sample alike add 2.0 ml of the ammonium phosphate solution.  The
         calibration standards must be prepared using the same type of acid and
         at the same concentration as will result in the sample to be analyzed
         after sample preparation.

Instrument Parameters (General)

     1.  Drying Time and Temp:  30 sec-125'C.
     2.  Ashing Time and Temp:  30 sec-500'C.
     3.  Atomizing Time and Temp:  10 sec-1900°C.
     4.  Purge Gas Atmosphere:  Argon
     5.  Wavelength:  228.8 nm
     6.  The operating parameters should be set as specified by the particular
         instrument manufacturer.
Notes
     1.  The above concentration values and instrument conditions are for a
         Perkln-Elmer HGA-2100, based on the use of a 20 ul injection, contin-
         uous flow purge ga's and non-pyrolytic graphite and are to be used as
         guidelines only.  Smaller size furnace devices or those employing
         faster rates of atomlzation can be operated using lower atomization
         temperatures for shorter time periods than the above recommended
         settings.

     2.  The use of background correction is required.

     3.  Contamination from the work area is critical in cadmium analysis.
         Use pipet tips which are free of cadmium.

     4.  For every sample analyzed, verification is necessary to determine
         that method of standard addition is not required (see Exhibit E).

     S.  If method of standard addition is required, follow the procedure given
         in Exhibit E.                 B-33

-------
                                     LEAD

           Method 239.2 CLP-M  (Atomic Absorption, furnace technique)


Optimum Concentration Range:   5-100 ug/1
Approximate Detection'Limit:   1 ug/1

Preparation of Standard Solution

     1.  Stock solution:  Carefully weigh  1.599 g of lead nitrate,  Pb  (N03)2
         (analytical reagent grade), and dissolve in deionlzed  distilled water.
         When solution is complete, acidify with 10 mL redistilled  HN<>3 and
         dilute to  1 liter with deionized  distilled water.  1  mL • 1 mg Pb
         (lOOOmg/L).

     2.  Lanthanum  Nitrate solution:  Dissolve 58.64 g of ACS reagent  grade
         La2<)3 la 100 ml cone. HN(>3 and dilute to 1000 ml with  deionized
         distilled  water.  1 ml -  50 mg La.

     3.  Working Lead solution:  Prepare dilutions of stock lead solution to  be
         used as calibration standards at  the time of analysis.  The calibration
         standards  must be prepared using  the same type  of  acid and at the same
         concentration as will result in the sample  to be analyzed  after sample
         preparation.  To each 100 ml of diluted standard add 10 ml of the
         lanthanum  nitrate solution.

Sample Preparation

     1.  To each 100 ml of prepared sample solution  add  10  ml of the lanthanum
         nitrate solution.

Instrument Parameters  (General)

     1.  Drying Time and Temp: 30 sec-125'C.

     2.  Ashing Time and Temp: 30 sec-500'C.

     3.  Atomizing  Time and Temp:   10 sec-2700'C.
     4.  Purge Gas  Atmosphere: Argon

     5.  Wavelength:   283.3 nm

     6.  Other operating parameters  should be set as specified  by the  particular
         instrument manufacturer.

Notes
      1.  The  above  concentration  values  and instrument  conditions are  for a
         Perkin-Elmer  HGA-2100,  based  on the use of  a 20 ul Injection, contin-
         uous flow  purge  gas  and  non-pyrolytic  graphite  and are to be  used as
          guidelines only.   Smaller size furnace devices  or those employing
          faster  rates  of  atomizatlon can be operated using lower acomizacion
          temperatures  for shorter time periods  than the above recommended .
          settings*

                                        B-34

-------
Method 239.2 CLP-M (cone.)
     2.  The use of background correction is required.

     3.  Greater sensitivity can be achieved using the 217.0 na  line, but
         the optimum concentration range is reduced.  The use of a lead
         electrodeless discharge lamp at this lower wavelength has been
         found to be advantageous.  Also a lower atomization temperature
         (24UO'C) may be preferred.

     4.  To suppress sulfate interference (up to 1500 ppm) lanthanum is
         added as the nitrate to both samples and calibration standards.
         (Atomic Absorption Newsletter Vol. K5, No. 3, p. 71, May-June 1976).

     5.  Since glassware contamination is a severe problem in lead analysis,
         all glassware*should .be cleaned immediately prior to use, and once
         cleaned, should not be open to the atmosphere except when necessary.

     6.  For every sample analyzed, verification is necessary to determine
         that method of standard addition is not required (see Exhibit E).

     7.  If method of standard addition is required, follow the procedure
         given in Exhibit E.
                                      B-35

-------
                                    SILVER

           Method 272.2 CLP-M (Atomic Absorption, furnace technique)


Optimum Concentration Range:  1-25 ug/1
Approximate Detection Limit:  0.2 ug/1

Preparation of Standard Solution

     1.  Stock solution:  Dissolve 1.575 g of AgN03 (analytical reagent
         grade) in deionized distilled water.  Add 10 mL of concentrated
         and make up to 1 liter.  1 mL » 1 mg Ag (1000 mg/L).

     2.  Prepare dilutions of Che stock solution to be used as calibration
         standards at the time of analysis.  These solutions are also to be
         used for "standard additions".

     3.  The calibration standards must be prepared using the same type of
         acid and at the same concentration as will result in the sample to
         be analyzed after sample preparation.

Instrument Parameters (General)

     1.  Drying Time and Temp:  30 sec-125'C.
     2.  Ashing Time and Temp:  30 sec-400*C.
     3.  Atomizing Time and Temp:  10 sec-2700'C.
     4.  Purge Gas Atmosphere:  Argon
     5.  Wavelength:  328.1 nm
     6.  Other operating parameters should be set as specified by the
         particular instrument manufacturer.

Notes
     1.  The above concentration values and instrument conditions are for
         a Perkin-Elmer HCA-2100, based on the use of a 20 ul injection,
         continuous flow purge gas and non-pyrolytic graphite and are to
         be used as guidelines only.  Smaller size furnace devices or those
         employing faster rates of atomization can be operated using lover
         atomization temperatures for shorter time periods than the above
         recommended settings.

     2.  The use of background correction is required.

     3.  The use of hallde acids should be avoided.

     4.  If absorption  to container walls or formation of AgCl is suspected,
         see NOTE 3 under the Direct Aspiration Method (Exhibit D, Attachment  4).

     5.  For every sample analyzed, verification is necessary to determine
         that method of standard addition is not required (see Exhibit  E).

     6.  If method of  standard addition is required,  follow  the procedure
         given  in Exhibit E.

                                       B-36

-------
                                  ATTACHMENT 3

                              Method  200.7 CLP-M*
        INDUCTIVELY COUPLtD PLASMA-ATOMIC EMISSION SPECTROMETRIC METHOD
                 FOR TRACE ELEMENT ANALYSIS OF WATER AND WASTES
1.   Scope and Application

     1.1   Dissolved elements are determined in filtered and acidified samples.
           Appropriate steps must be taken in all analyses to ensure that
           potential interferences are taken into account.  This is especially
           true when dissolved solids exceed 150U mg/L.  (See 5.)

     1.2   Total elements are determined after appropriate digestion procedures
           are performed.  Since digestion, techniques increase the dissolved
           solids content of the samples, appropriate steps must be taken to
           correct for potential interference effects.  (See 5.)

     1.3*   Table 1 lists elements along with recommended wavelengths and typical
           estimated instrumental detection limits using .conventional pneumatic
           nebulizatlon.  Actual working detected limits are sample dependent
           and as the sample matrix varies, these concentrations may also vary.
           In time, other elements maybe added as more information becomes
           available and as required.

     1.4   Because of the differences between various makes and models of
           satisfactory instruments, no detailed instrumental operating
           instructions can be provided.  Instead, the analyst is referred
           to the Instructions provided by the manufacturer of the particular
           instrument.

2.   Summary of Method

     2.1   The method describes a technique for the simultaneous or sequential
           multielement determination of trace elements in solution.  The basis
           of the method is the measurement of atomic emission by an optical
           spectroscopic technique.  Samples are nebulized and the aerosol
           that is produced is transported to the plasma torch where excitation
           occurs.  Characteristic atomic-line emission spectra are produced by
           a radio-frequency inductively coupled plasma (ICP).  The spectra are
           dispersed  by a grating spectrometer and  the intensities of the line
           are monitored by photomultiplier tubes.  The photocurrents from Che
           photomultiplier  tubes are.processed and  controlled by a computer
           system.  A background correction technique is required co compensate
           for variable  background contribution  to  the determination of  trace
           elements.  Background must be measured adjacent to analyte lines on
 *CL?-M Modified for the  Contract  Laboratory Program
                                      B-37

-------
Method 200.7 CLP-M (cont.)


           samples during analysis.  The position selected  for  the  background
           intensity measurement, on either or  both  sides of  the  analytical
           line, will be determined by the complexity  of the  spectrum adjacent
           to the analyte line.  The position used'must  be  free of  spectral
           interference and reflect the same change  in background intensity  as
           occurs at the analyte wavelength measured.  Background correction is
           not required in cases of line broadening  where a background correc-
           tion measurement would actually degrade the analytical result.  The
           possibility of additional interferences named In S.I (and  tests for
           their presence as described In 5.2)  should  also  be recognized and
           appropriate corrections made.

3.   Definitions

     3.1   Dissolved — Those elements which-will pass through  a  0.45 urn
           membrane filter.

     3.2   Suspended — Those elements which are retained by  a  0.45 urn
           membrane filter.

     3.3   Total — The concentration determined on  an unfiltered sample
           following vigorous digestion.

     3.4   Instrumental detection limits — See Exhibit E,  pages  2-4.

     3.5   Sensitivity — The slope of the analytical  curve,  i.e. functional
           relationship between emission intensity and concentration.

     3.6   Instrument cheek standard — A multielement standard of  known
           concentrations prepared by the analyst to monitor  and  verify
           Instrument performance on a dally basis.  (See 7.6.1.)

     3.7   Interference cheek sample — A solution containing both  interfering
           and analyte elements of known concentration that can be  used co
           verify background and interelement correction factors.  (See 7.6.2.)

     3.0   Quality control sample — A solution obtained from an  oucslde source
           having known concentration values to be used to  verify the calibra-
           tion standards.  (See 7.6.3.)

     3.9   Calibration standards — A series of known  standard  solutions used
           by the analyst for calibration of the instrument (i.e.,  preparation
           of the analytical curve).  (See 7.4.)

     3.10 Linear dynamic range — The concentration range  over which the
           analytical curve remains linear as determined in Exhibit E.

     3.11 Reagent blank — A volume of deionlzed, distilled  water  containing
           the same acid matrix as the calibration standards  carried  through
           the entire analycical scheme.  (See  7.5.2.)
                                     B-38

-------
Method 200.7 CLP-M (cone.)


     3.12  Calibration blank — A volume of deionized,  distilled  water acidified
           vlth HN03 and UC1.  (See  7.5.1.)

     3.13  Method of standard addition — The standard  addition technique
           Involves the use of the unknown and  the  unknown-^plus-a-known amount
           of standard by adding known amounts  of standard  to  one or more
           allquots of the processed sample solution.

4.   Safety

     4.1   The toxlcity or carcinogenicity of each  reagent  used In this method
           has not been precisely defined; however, each chemical compound
           should be treated as a potential health  hazard.   The laboratory
           is responsible for maintaining a current awareness  file of OSHA
           regulations regarding the safe handling  of  the chemicals specified
           in this method.  A reference  file  of material handling data sheets
           should also be made available to all personnel involved in the
           chemical analysis.  Additional references to laboratory safety  are
           available and have been identified  (11.7, 11.8 and  11.9) for the
           information of the analyst.

5.   Interferences

     5.1   Several types of  interference effects may contribute  to in-
           accuracies  in the  determination  of trace elements.   They can be
           summarized  as follows:

            5.1.1   Spectral  interferences can be categorized as  1) overlap of
                   a  spectral line  from another element; 2) unresolved overlap
                   of  molecular  band spectra;  3) background contribution from
                    continuous or recombination -phenomena; and 4)  background
                    contribution  from stray  light from the line emission of high
                    concentration elements.   The first of these effects can be
                    compensated  by  utilizing a computer correction of the raw
                    data,  requiring the monitoring and measurement of the inter-
                    fering  element.   The second effect may require selection of
                    an alternate wavelength.   The third and fourth effects can
                    usually be compensated by a background correction adjacent
                    to the  analyte line.  In addition, users of simultaneous
                    multi-element instrumentation must assume the responsibility
                    of verifying the absence of spectral interference from an
                    element that could occur in a sample but for which there is
                    no channel in the instrument array.  Listed in Table 2 are
                    some Interference effects for the recommended wavelengths
                    given in Table 1.  The data in Table 2 are intended for use
                    only as a rudimentary guide for' the indication of potential
                    spectral interferences.  For this purpose, linear relations
                    between concentration and intensity for the analytes and the
                    interferents can be assumed.  The interference  information,
                    which was collected at the Ames  Laboratory1,  is  expressed as
 'Ames Laboratory, USUOE, Iowa State University, Ames Iowa 50011
                                       B-39

-------
Method 200.7 CLP-M (cone.)
                   analyte concentration equivalents  (i.e.  false  analyte  concen-
                   trations) arising from 100 mg/L of the interferenc  element.
                   The suggested use of this information is 'as  follows:   Assume
                   that arsenic (at 193.696 mn) is to be determined  in a  sample
                   containing approximately 10 mg/L of of aluminum.  According
                   to Table 2, 100 mg/L of aluminum would yield a false signal
                   for arsenic equivalent to approximately  1.3  mg/L.   Therefore,
                   10 mg/L of aluminum would result in a false  signal  for arsenic
                   equivalent to approximately 0.13 mg/L.   The  reader  is  cautioned
                   that other analytical systems may  exhibit  somewhat  different
                   levels of interference Chan those  shown  in Table  2, and that
                   the Interference effects must be evaluated for each individual
                   system.  Only those interferents listed  were investigated  and
                   the blank spaces in Table 2 indicate that  measurable inter-
                   ferences were not observed from the interferent concentrations
                   listed in Table 3.  Generally, interferences were discernible
                   if they produced peaks or background shifts  corresponding  to
                   2-5Z of the peaks generated by the analyte concentrations  also
                   listed in Table 3.

                   At present, information on the listed silver and  potassium
                   wavelengths are not available but  it has been  reported that
                   second order energy from the magnesium 383.231 nm wavelength
                   Interferes with the listed potassium line  at 766.491 an.

            5.1.2   Physical interferences are generally considered to  be  effects
                   associated with the sample nebulization  and  transport  proc-
                   esses.  Such properties as change  in viscosity and  surface
                   tension can cause significant inaccuracies especially  in
                   samples which may contain high dissolved solids and/or acid
                   concentrations.  The use of a peristaltic pump may  lessen
                   these Interferences.  If these types of  interferences  are
                   operative,  they must be reduced  by dilution  of the  sample
                   and/ or utilization of standard addition techniques.   Another
                   problem which  can occur from high  dissolved  solids  is  sale
                   buildup at  the tip of the nebulizer.  This affects  aerosol
                   flow race  causing instrumental drift.  Vetting Che  argon
                   prior to nebulization, the use of  a tip  washer, or  sample
                   dilution have  been used to  control this  problem.  Also, it
                   has been reported that better control of the argon  flow race
                   improves instrument  performance.   This  is accomplished with
                   the use of  mass flow controllers.

            5.1.3  Chemical  interferences  are  characterized by  molecular  conr-
                   pound  formation,  lonlzation effects and  solute vaporization
                   effects,   formally  these  effects  are  not pronounced with the
                    ICP  technique, however, if  observed  they can be minimized  by
                    careful  selection of  operating  conditions (that is, incldenc
                   power,  observation  position,  and  so  forth),  by buffering of
                    the  sample,  by matrix  matching,  and  by  standard addition
                    procedures.   These  types  of  interferences can  be  highly
                    dependent  on matrix type  and  the specific analyce element.
                                      B-40

-------
Method 200.7 CLP-M (cone.)


     5.2   For each group of samples of a similar matrix type and concentration
           (i.e., low, medium) for each Case of samples, or for each  20  samples
           received, whichever is more frequent, the following tests  must be
           performed prior to reporting concentration data for analyte elements.

           5.2.1   Serial dilution— If the analyte concentration is suffi-
                   ciently high (minimally a factor of  10 above the instrument
                   detection limit after dilution), an  analysis of a  1:4 dilution
                   must agree within 10 percent of the  original determination.
                   Serial dilution results must be reported on QC Report Form IX.
                   Samples identified as Field Blanks cannot be used  for serial
                   dilution analysis.

                   If the dilution analysis is not within 10Z, a chemical or
                   physical interference effect should  be suspected,  and the
                   data oust be flagged with an T.

6.   Apparatus

     6.1   Inductively Coupled Plasma-Atomic Emission Spectrometer.

           6.1.1   Computer controlled atomic emission  spectrometer with back-
                   ground correction.

           6.1.2   Radiofrequency generator.

           6.1.3   Argon gas supply, welding grade or better.

     6.2   Operating conditions — Because of the differences between various
           makes and models of satisfactory instruments, no detailed  operating
           instructions can be provided.  Instead, .the  analyst should follow the
           instructions provided by the manufacturer of the particular instrument.
           Sensitivity, Instrumental detection limit, precision, linear  dynamic
           range, and interference effects oust be investigated and established
           for each individual analyte line on that particular Instrument.  All
           measurements must be within the instrument linear range where correc-
           tion factors are valid.  It is the responsibility of the analyst to
           verify that the instrument configuration and operating conditions
           used satisfy the analytical requirements and to maintain quality con-
           trol data confirming instrument performance  and analytical results.

7.   Reagents and standards

     7.1   Acids used in the preparation of standards and for sample  processing
           must be ultra-high purity grade or equivalent.  Redistilled acids
           are acceptable.

           7.1.1   Acetic acid, cone, (sp gr 1.06).

           7.1.2   Hydrochloric acid, cone, (sp gr  1.19).

           7.1.3   Hydrochloric acid. (1+1):  Add 500 mL cone. HC1  (sp gr  1.19)
                    to 400 mL delonlzed, distilled water and dilute  Co L  liter.
                                      ' B-41

-------
Method 200.7 CLP-M  (cone.)


         7.1.4   Nitric acid, cone,  (sp gr  1.41).

         7.1.5   Nitric acid. (1+1):  Add 500 mL  cone.  HN03  (sp gr'1.41)  to
                 400 mL deionized, distilled water  and  dilute to 1  liter.

   7.2   Deionized, distilled water;  Prepare by  passing  distilled  water
         through a  mixed  bed of  cation and  anion  exchange resins.   Use
         deionized, distilled water  for the preparation of all reagents,
         calibration standards and as dilution water.   The purity of  this
         water must be equivalent to ASTM Type II reagent water of  Specifi-
         cation D 1193 (14.t>).

   7.3   Standard stock solutions may be purchased  or prepared from ultra
         high purity grade  chemicals or metals.   All salts must be  dried  for
         1 h at 105° unless- otherwise specified.

         (CAUTION:  Many  metal salts are extremely  toxic  and may be fatal  if
         swallowed.  Wash hands  thoroughly  after  handling.)  Typical stock
         solution preparation procedures follow:

         7.3.1   Aluminum solution,  stock.  1 mL • 100 ug  Al:   Dissolved 0.100
                 g  of aluminum metal in an  acid mixture of 4 mL of  (1+1) HC1
                 and 1 mL of cone. HN03 in  a beaker.  Warm gently to  effect
                 solution.  When solution is complete,  transfer quantitatively
                 to a liter flask, add an additional 10 mL of (1+1) HC1 and
                 dilute to  1000  mL with deionized,  distilled water.

         7.3.2   Antimony solution stock, 1 oL -  100 ug Sb:   Dissolve 0.2669
                 g  K(SbO)C4h406  in deionized distilled  water, add 10  mL (1+1)
                 ECl and  dilute  to 1000 mL  with deionized, distilled  water.

         7.3.3   Arsenic  solution, stock. 1 mL •  100 ug As:   Dissolve 0.1320
                 g  of As203 in 100 mL of deionized,  distilled water containing
                 0.4 g NaOH.  Acidify the solution  with 2 mL cone.  HNC>3 and
                 dilute to  1,000 mL  with deionized,  distilled water.

         7.3.4   Barium solution, stock. 1  mL - 100 ug  Ba:   Dissolve  0.1516  g
                 Bad2 (dried at 250°C for  2 hrs)-in 10 mL deionized, distilled
                 water with 1 mL (1+1) UC1.  Add  10.0 mL  (1+1) HC1  and dilute
                 to 1,000 mL with deionized, distilled  water.

         7.3.5   Beryllium  solution, stock, 1 mL  •  100  ug Be:  Do not dry.
                 Dissolve 1.966  g BeSO4'4H20, in  deionized,  distilled water,
                 add 10.0 mL cone. UN03 and dilute  to 1,000  mL with deionized,
                 distilled  water.

         7.3.6   Boron solution, stock, 1 mL • 100  ug B:  Do not dry. Dissolve
                 0.571b g anhydrous  (13803 in deionized, distilled water and
                 dilute to  1,000 nL. Use a reagent meeting  ACS specifications
                 keep  the bottle tightly stoppered  and  store in a desiccator
                 to prevent the  entrance of atmospheric moisture.
                                     B-42

-------
Method 200.7 CLP-M (cone.)
           7.3.7   Cadmium solution, stock.  1 mL -  100 ug  Cd:   Dissolve  0.1142
                   g CdO in a minimum amount of  (1+1) HliC^.   Heat  to increase
                   rate of dissolution.  Add 10.0 mL cone. HN03 and  dilute to
                   1,000 mL with deionized,  distilled water.

           7.3.8   Calcium solution, stock.  1 mL -  100 ug  Ca:   Suspend 0.2498  g
                   CaO>3 dried  at  IbO'C  for  1 h  before weighing in deionized,
                   distilled water and dissolve cautiously with a  minimum amount
                   of  (1+1) HNC>3.  Add 10.0  mL cone. HN(>3  and dilute to  1,000  mL
                   with deionized, distilled water.

           7.3.9   Chromium solution, stock. 1 mL • 100  ug Cr:   Dissolve 0.1923
                   g of Cr03 in deionized, distilled water.   When  solution is
                   complete acidify with 10  mL cone. H1K>3  and dilute to  1,000  mL
                   with deionized, distilled water.

           7.3.10  Cobalt solution stock.  1  mL • 10 ug Co: Dissolve 0.1000 g
                   of  cobalt metal in a  minimum  amount of  (1+1) HN03. Add 10.0
                   mL  (1+1) UC1 and dilute to  1,000 mL with deionized, distilled
                   water.

           7.3.11  Copper solution, stock. 1 mL  • 100 ug Cu:   Dissolve 0.1252
                   g CuO in a minimum amount of  (1+1) UM03*   Add 10.0 mL cone.
                   HN03 and dilute to 1,000  mL with deionized.  distilled water.

           7.3.12  Iron solution,  stock, 1 mL •  100 ug Fe: Dissolve 0.1430 g
                   Fe2<>3 in a warm mixture of  20 mL (1+1)  UC1 and  2  mL of cone.
                   HN03*  Cool, add an additional 5 mL of  cone. EN03 and dilute
                   to  1,000 mL  with deionized, distilled water.
            7.3.13  Lead solution,  stock.  1 mL - 100 ug Pb:   Dissolve 0.1599 g
                    Pb(N03)2 in a minimum  amount of  (1+1)  HN03.   Add 10.0 mL of
                    cone.  HN03 and dilute  to 1,000 mL with deionized, distilled
                    water.

            7.3.14  Magnesium solution,  stock, 1 mL  - 100  ug Mg:   Dissolve 0.1658
                    g MgO in a minimum amount of (1+1) HM^.  Add 10.0 mL cone.
                    HN03 and dilute to 1,000 mL with deionized,  distilled water.


            7.3.15  Manganese solution,  stock. 1 mL  - 100  ug Mn:   Dissolve O.lOOO
                    g of manganese metal in the acid mixture, 10 nL cone.  HC1
                    and 1 mL epnc. UNO3, and dilute  to 1,000 mL with deionized,
                    distilled water.

            7.3.16  Molybdenum solution, stock, 1 mL • 100 ug Mo:  Dissolve
                    0.2043 g (NH4)2Mo04 in deionized, distilled water and dilute
                    to 1,000 mL.

                                      B-43

-------
Method 200.7 CLP-M  (cone.)
           7.3.17  Nickel solution,  stock.  1 mL  -  100  ug  Ni:   Dissolve 0.1000 g
                   of nickel metal in  10  nL hot  cone*  HN03,  cool and dilute co
                   1,000 nL with  deionized, distilled  water.

           7.3.1S  Potassium solution,  stock.  1  nL - 100  ug  K:   Dissolve 0.1907
                   g KC1, dried at 110'C,  in deionized, distilled water.  Dilute
                   to 1,000 mL.

           7.3.19  Selenium solution,  stock. 1 mL  • 100 ug Se:   Do not dry.
                   Dissolve 0.1727 g H2Se°3 (actual assay 94.6Z) in deionized,
                   distilled water and dilute  co 1,000 mL.
            7.3.20   Silica  solution,  stock,  1 mL • 100 ug S102:   Do not dry.
                    Dissolve  0.4730  g Na2Si03'9H20 in deionized,  distilled water.
                    Add  10.0  mL  cone.  BNOj and  dilute to 1,000 mL with deionized,
                    distilled water.

            7.3.21   Silver  solution,  stock.  1 mL - 100 ug Ag:  Dissolve 0.1575 g
                    AgN(>3 in  100 mL  of deionized, distilled water and 1U mL cone.
                    HN<>3.   Dilute to  1,000 mL with deionized,  distilled water.

            7.3.22   Sodium  solution,  stock.  1 mL « 100 ug Na:  Dissolve 0.2542 g
                    MaCl in deionized, distilled water.  Add 10.0 mL cone. HN03
                    and  dilute to 1,000 mL .with deionized, distilled water.

            7.3.23   Thallium  solution, stock.  1 mL - 100 ug Tl:   Dissolve 0.1303
                    g  T1N03 in deionized,  distilled water.  Add  10.0 mL cone.
                    HNOj and  dilute  to 1,000 mL with deionized,  distilled water.

            7.3.24   Vanadium  solution, stock.  1 mL - 100 ug V:  Dissolve 0.2297
                    NH4V03  in a  minimum amount  of cone. HN03*  Heat to Increase
                    rate of dissolution.  Add  10.0 mL cone. HN03 and dilute to
                    1,000 mL  with deionized, distilled water.

            7.3.25   Zinc solution, stock.  1  mL  - 100 ug Zn:  Dissolve 0.1245 g
                    ZnO  in  a  minimum amount  of  dilute HM03.  Add 10.0 mL cone.
                    HM03 and  dilute  to 1,000 mL with deionized,  distilled water.

      7.4   Mixed calibration standard solutions — Prepare mixed calibration
            standard solutions by combining  appropriate volumes  of the stock
            solutions  in volumetric  flasks.   (See 7.4.1 thru 7.4.5.)  Add 2
            mL of (1+1)  UN03  and 1U  mL of  (1+1) HC1 and dilute to 100 mL with
            deionized. distilled water.  (See  Notes 1 and 6.)   Prior to pre-
            paring the mixed  standards, each stock solution should be analyzed
            separately to determine  possible spectral interference or the
            presence of  Impurities.   Care  should be taken when preparing che
            mixed standards that the elements  are compatible and stable.
            Transfer the mixed standard solutions to a FEP fluorocarbon or
            unused polyethylene  bottle for storage.  Fresh mixed scandards
            should be prepared as needed with  the realization that concencraclon

                                       B-44

-------
Method 20U.7 CLP-M (cone.)


           can change on aging.  Calibration standards must be  initially
           verified using a quality  control sample  and monitored  weekly for
           stability (see 7.6.3).  Although not specifically  required, some
           typical calibration standard  combinations  follow when  using those
           specific wavelengths listed in Table 1.

           7.4.1   Mixed standard  solution I — Manganese, beryllium,  cadmium,
                   lead, and zinc.

           7.4.2   Mixed standard  solution II —  Barium,  copper,  iron, vanadium,
                   and  cobalt.

           7.4.3   Mixed standard  solution III — Molybdenum, silica,
                   arsenic, and selenium.

           7.4.4   Mixed standard  solution IV —  Calcium, sodium, potassium,
                   aluminum, chromium and nickel.

           7.4.5   Mixed standard  solution V — Antimony, boron,  magnesium,
                   silver,  and  thallium.

                        NOTE 1:  If  the  addition  of silver to the recommended
                        acid pombination results  in an  initial  precipitation
                        add 15  mL  of deionized distilled  water  and warn  the
                        flask until  the  solution  clears.  .Cool  and dilute to
                        100 mL  with  deionized,' distilled  water.  For this
                        acid combination the  silver concentration should  be
                        limited to 2 mg/L.  Silver  under  these  conditions, is
                        stable  in  a tap  water matrix  for  30  days.  Higher
                        concentrations of silver  require  additional HC1.

      7.5  Two types  of blanks  are required  for  the analysis.  The calibration
           blank  (3.13) is  used in establishing  the analytical  curve while the
            reagent blank  (preparation blank,  3.12)  is .used to correct for
           possible  contamination  resulting from varying  amounts  of the  acids
            used in the sample  processing.
            7.5.1   The calibration blank is prepared by diluting 2 mL of
                    UMU3 and 10 mL of (1*1) HC1 to 100 mL with deionized,
                    distilled water.  (See Note 6.)  Prepare a sufficient
                    quantity to be used to flush the system -between standards
                    and samples.

            7.5.2   The reagent blank (or preparation blank - See Exhibit E)
                    must contain all the reagents and in the same volumes as
                    used in the processing of the samples.  The reagent blank
                    must be carried through the complete procedure and contain
                    the same acid concentration in the final solution as the
                    sample solution used for analysis.

                                      B-45

-------
Method 2CU.7 CLP-M (cone.)


     7.6   In addition the calibration standards, an instrument  check  standard
           (3.6), an interference check sample  (3.7) and  a quality  control
           sample (3.8) are also required for Che analyses.

           7.6.1   The instrument check standard for continuing  calibration
                   verification is prepared by  the analyst by  combining  com-
                   patible elements at a concentration  equivalent to the mid-
                   point of their respective calibration  curves.  (See 10.1.3.)

           7.0.2   The interference check sample is prepared by  the analyst,
                   or obtained from EPA if available (Exhibit  E).

           7.6.3   The quality control sample for the initial  calibration
                   verification should be prepared in the same acid matrix
                   as the calibration standards and in  accordance with the
                   Instructions provided by the supplier. EPA will either
                   supply a quality control sample or information where  one
                   of equal quality can be procured.  (See 10.1.1.)

8.  Procedure

     8.1   Set up Instrument with proper operating parameters  established in
           Section 6.2.  The instrument must be allowed to- become thermally
           stable before beginning.  This usually requires at  least 30 min.
           of operation prior to calibration.

     8.2   Initiate appropriate operating configuration of computer.

     8.3   Profile and calibrate Instrument according to  instrument manufac-
           turer's recommended procedures, using mixed  calibration  standard
           solutions such as those described in Section 7.4.   Flush the  system
           with the calibration blank (7.5.1) between each standard.   (See
           NOTE 7.) (Use the average intensity  of multiple exposures for both
           standardization and sample analysis  to reduce  random  error.)

           NOTE 7:  For boron concentrations greater than 500  ug/L  extended
           flush  times of 1 to 2 minutes may be required.

     8.4   Begin  the sample run flushing the system with  the calibration blank
           solution  (7.5.1) betveen each sample.   (See  NOTE  7.)  Analyze the
           instrument check standard  (7.6.1) and the calibration blank (7.5.1)
           each  10 samples.

9.   Calculation

     9.1   Reagent blanks (preparation blanks)  should be  treated as specified
            in Exhibit E.

     9.2    If dilutions were performed, the  appropriate factor muse be applied
            to sample  values.

      9.3    Data oust  be  reported  in ug/L.
                                       B-46

-------
Method 200.7 CLP-M  (cone.)
10.  Quality Control  (Instrumental)

     10.1  Check the  instrument standardization by analyzing appropriate
           quality control check standards as  follows:

           10.1.1  A  quality control sample  (7.6.3) must be used  daily  for  the
                   initial calibration verification (See Exhibit  E).  A fresh
                   dilution of this sample shall be analyzed every week there-
                   after to monitor their stability.   If the results  are  not
                   within *10Z of the true value listed for the control sample,
                   prepare a new calibration standard and  recalibrate the
                   instrument.  If this does not correct the problem, prepare a
                   new  stock standard and a  new calibration standard  and  repeat
                   the  calibration.

           10.1.2  Analyze the calibration blank (7.5.1) at a  frequency of  10Z.
                   The  result should be within ± contract  required detection
                   levels (Exhibit C).  If the result is not within the control
                   level, terminate the analysis, correct  the  problem and
                   recalibrate the instrument  (See Exhibit E).


           10.1.3  For  continuing calibration  verification, analyze an  appro-
                   priate instrument check standard (7.6.1) containing  the
                   elements of interest at a frequency of  10Z.  This  check
                   standard is used to determine instrument drift.  If  agree-
                   ment is not within ±102 of  the expected values, the  analysis
                   is out of control.  The analysis must be terminated, the
                   problem corrected, the instrument  recalibrated, and  the
                   preceding 10 samples reanalyzed  (See Exhibit E).

           10.1.4  To verify interelement and  background correction factors
                   analyze the ICP interference check sample (7.6.2)  at the
                   beginning, and end of the sample run or a nimimum  of twice
                   per  8 hour work shift whichever  is more frequent.  The check
                   sample must be analyzed initially  at least  5  times repeti-
                   tively to establish a mean  value and standard  deviation.
                   Results must fall within  the established control limits. If
                   not, terminate the analysis, correct the problem,  recalibrate
                   the  instrument, and reanalyze the  samples  (See Exhibit E).

 11.  Bibliography

     1.  Winge,  R. K.,  V.J. Peterson, and V.A. Tassel,  "Inductively Coupled
         Plasma-Atomic  Emission Spectroscopy Prominent Lines," EPA-600/4-79-017.

     2.  Winefordner, J.D.,  "Trace Analysis:  Spectroacopic Methods for Elements,
         Chemical  Analysis, Vol.  46, pp. 41-'42.
                                        B-47

-------
Method 200.7 CLP-M (cont.)


     3.  Handbook for Analytical Quality Control in Water and Wastevater
         Laboratories, EPA-600/4-79-019.

     4.  Garbarino, J.R. and Taylor, H.E., "An Inductively-Coupled Plasma
         Atomic Emission Spectrometric Method for Routine Water Quality
         Testing," Applied Spectroscopy .3p_i No< 3(1979).

     5.  "Methods for Chemical Analysis of Water and Wastes," EPA-600/4-79-020.

     6.  Annual Book of ASTM Standards, Part 31.

     7.  "Carcinogens - Working With Carcinogens," Department'of Health,
         Education, and Welfare, Public Health Service, Center for Disease
         Control, National Institute for Occupational Safety and Health,
         Publication No. 77-206, Aug. 1977.

     8.  "OSHA Safety and Health Standards, General Industry," (29 CFR 1910),
         Occupational Safety and Health Administration, OSHA 2206, (Revised,
         January 1976),
                      .
     9.  "Safety in Academic Chemistry Laboratories, American Chemical Society
         Publications, Committee on Chemical Safety, 3rd Edition, 1979.

    10.  "Inductively Coupled Plasma-Atomic Emission. Spectrometric Method of
         Trace Elements Analysis of Water and Waste", Method 200.7 modified
         by CLP Inorganic Data/Protocol Review Committee; original method by
         Theodore D. Martin, EMSL/Cinclnnati.
                                      B-48

-------
Method 200.7 CLP-M (cone.)
Element
               TABLE 1 - RECOMMENDED WAVELENGTHS<2) AND ESTIMATED
                         INSTRUMENTAL DETECTION  LIMITS
Wavelength,
Estimated Detection
  Limit, ug/L<2)
Aluminum
Antimony
Arsenic
Barium
Beryllium
Boron
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Molybdenum
Nickel
Potassium
Selenium
Silica (Si02)
Silver
Sodium
Thallium
Vanadium
Zinc
308.215
206.833
193.696
455.403
313.042
249.773
226.502
317.933
267.716
228.616
324.754
259.940
220.353
279.079
257.610
202.030
231.604
766.491
196.026
288.158
328.068
588.995
190.864
292.402
213.856
45
32
53
2
0.3
5
4
10
7
7
6
7
42
30
2
8
15
seeO)
75
58
7
29
40
8
2
     The  wavelengths  listed are recommended because of  their sensitivity and
     overall  acceptance.   Other wavelength may be substituted if they can
     provide  the needed sensitivity and are treated with the same corrective
     techniques  for spectral interference.  (See 5.1.1). The use of alternate
     wavelengths must be 'reported (in nm) with the sample data.

     The  estimated instrumental detection limits as shown are taken from
     "Inductively Coupled  Plasma-Atomic Emission Spectroscopy-Promlnent Lines,
     EPA-600/4-79-017.  They are given as a guide for an instrumental Hole.
     The  actual  method detection limits are sample dependent and may vary as
     the  sample  matrix varies.

     highly dependent on operating conditions and plasma position.
                                      B-49

-------
TABLE 2.  EXAMPLE OP ANALYTE CONCENTRATION EQUIVALENTS  (ng/L) ARISING  FROM
                    INTEKPEKENTS AT THE  100 ng/L LEVEL
Analyte
Aluminum
Antimony
Araenlc
Barium
Beryllium
Boron
Cadmium
„ Calcium
CO
' Chromium
0
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Molybdenum
Nickel
Selenium
Silicon
Sodium
Thallium
Vanadium
Zinc
Wavelength,
nm
308.215
206.833
193.696
455.403
313.042
249.773
226.502
317.933
267.716

228.616
324.754
259.940
220.353
279.079
257.610
202.030
231.604
196.026
288.158
•588.995
190.864
292.402
2IJ.H56

Al Ca
w 	 a
0.47
1.3
__ __
—
0.04
__ —
— —
— —

__ — .
—
__ __
0.17
0.02
0.005 —
0.05
— —
0.23
__ __
—
0.30
__ _ _
— — — —
Interferent
Cr Cu Pe Mg Hn Nl
— 0.21
2.9 — 0.08 — —
0.44
— __ __ — — — —
— — — — —
_ n 11 __ — — —
__ _— U.Jfc
— — 0.03 — — 0.02
0.08 — 0.01 0.01 0.04 —
— o.003 — 0.04

0.03 — 0.005 — — 0.03
0.003 — — —
0.12
__ — — — — —
0.11 — 0.13 — 0.25 —
0.01 — 0.002 0.002
0.03 —
— — — — — — — — — — — —
0.09
0.07
— — — —
— — ~~ •"— ~*~ ~~ ~~
0.05 -- 0.005
— 0.14 -- -- — 0.29

Ti
__
.25
~~
—
0.04
~
—
0.03
—

0.15
0.05
^^
~
0.07
^~»
—
— —
— —
—
0.08
— —
0.02
™~

V
1.4
0.45
1.1 •
—
0.05
""
—
0.03
0.04

—
0.02
""
—
0.12
^^
—
—
~
0.01
—
^"™
—
™~

-------
Method 200.7 CLP-M (cone)
      TABLE 3.  INTERFERENT AND ANALYTE ELEMENTAL CONCENTRATIONS USED
                 FOR INTERFERENCE MEASUREMENTS IN TABLE 2  (EXHIBIT 0)
Analytes
Al
As
B
Ba
Be
Ca
Cd
Co
Cr
Cu
Fe
Mg
Mn
Mo'
Na
Nl
Pb
Sb
Se
Si
Tl
V
Zn
(mg/L)
10
10
10
1
1
1
10
1
1
1
1
1
1
10
10
10
10'
10
10
1
10
1
10
Interferents
Al
Ca
Cr
Cu
Fe
Mg
Mn
Ni
Tl
V













(mg/L)
1000
1000
200
200
1000
1000
200
200
200
200













                                      B-51

-------
                                  ATTACHMENT 5

                                    MERCURY

               Method 245.1 CLF-M* (Manual Cold Vapor Technique)


1.   Scope and Application

     1.1   In addition to inorganic forms of mercury, organic mercurials
           may also be present.  These organo-mercury compounds will not
           respond to the cold vapor atomic absorption technique unless they
           are first broken down and converted to mercuric ions.  Potassium
           permanganate oxidizes many of these compounds, but recent studies
           have shown that a number of organic mercurials, including phenyl
           mercuric acetate and methyl mercuric chloride, are only partially
           oxidized by this reagent.  Potassium, persulfate has been found to
           give approximately 100Z recovery when used as the oxldant with
           these compounds.  Therefore, a persulfate oxidation step following
           the addition of the permanganate has been included to insure thac
           organo-mercury compounds, if present, will be oxidized to the
           mercuric ion before measurement.  A heat step is required for
           methyl mercuric chloride when present in or spiked to a natural
           system.  For distilled water the heat step is not necessary.

     1.2   The range of the method may be varied through instrument and/or
           recorder expansion.  Using a 100 ml sample, a detection limit of
           0.2 ug Hg/1 can be achieved (see Appendix 11.2).

2.   Summary of Method

     2.1   The flameless AA procedure is a physical method based on the
           absorption of radiation at 253.7 nm by mercury vapor.  Organic
           mercury compounds are oxidized and the mercury is reduced to the
           elemental state and aerated from solution in a closed system.
           The mercury vapor passes through a cell positioned in the light
           path of an atomic absorption spectrophotomecer.  Absorbance  (peak
           height) la measured as a function of mercury concentration and
           recorded  in the usual manner.
            *CLP-M modified for the Contract  Laboratory Program


                                       B-52

-------
Method 245.1 CLF-H  (cone.)
3.   Sample Handling and Preservation

     3.1   Until more conclusive data are obtained,  samples  should  be
           preserved by acidification with nitric  acid  to  a  pH of 2 or
           lover Immediately at the time of collection.

4.   Interference

     4.1   Possible lnterferen.ce from sulflde Is eliminated  by the  addition
           of potassium permanganate.  Concentrations as high  as 20 mg/1  of
           sulflde as sodium sulflde do not Interfere with the recovery of
           added inorganic mercury from distilled  water.

     4.2   Copper has also been reported to Interfere;  however, copper
           concentrations as high as 10 mg/1 had no  effect on  recovery of
           mercury from spiked samples.

     4.3   Sea waters, brines and industrial effluents  high  in chlorides
           require additional permanganate (as much  as  25  ml).  During the
           oxidation step, chlorides are converted to free chlorine  which
           will also absorb radiation of 253 nm.   Care  must  be taken to
           assure that free chlorine is absent before the  mercury is reduced
           and swept into the cell.- This may be accomplished  by using an
           excess of hydroxylamlne sulfate reagent (25  ml).  In addition,
           the dead air space in the BOD bottle must be purged before  the
           addition of stannous sulfate.  Both inorganic and organic mercury
           spikes have been quantitatively recovered from  the  sea water using
           this technique.

     4.4   Interference from certain volatile organic materials which  will
           absorb at this wavelength is also possible.  A  preliminary  run
           without reagents should determine if this type  of interference is
           present (see Appendix 11.1).

5.   Apparatus

     5.1   Atomic Absorption Spectrophotometer:  (See Note 1)  Any atomic
           absorption unit having an open sample presentation  area  in  which
           to mount the absorption cell is suitable.  Instrument settings
           recommended by the particular manufacturer should be followed.
           NOTE !:•  Instruments designed specifically for  the  measurement
           of mercury using the cold vapor technique are commercially  available
           and may be substituted for the atomic absorption  spectrophotometer.
                                        B-53

-------
Method 245.1 CLP-M (cone.)


     5.2   Mercury Hollow Cathode Lamp:  Vestinghouse WL-22847,  argon  filled,
           or equivalent.

     5.3   Recorder:  Any multi-range variable speed recorder  that  is  compatible
           'with the UV detection system is suitable.

     5.4   Absorption Cell:  Standard spectrophotometer cells  10 cm long, having
           quartz end windows may be used.  Suitable cells may be constructed
           from plexiglass tubing, 1" O.D. Z 4-1/2".  The ends are  ground per-
           pendicular to the longitudinal axis and quartz windows (1"  diameter
           Z 1/16" thickness) are cemented in place.  The cell is strapped  to
           a burner for support and aligned in the light beam  by use of  two 2"
           by 2" cards.  One Inch diameter holes are cut in the  middle of each
           card; Che cards are than placed over each end of the  cell.  The  cell
           is then positioned and adjusted vertically and horizontally to find
           the maximum transmlttance.

     5.5   Air Pump:  Any peristaltic pump capable of delivering 1  liter of air
           per minute may be used.  A Masterflex pump with electronic  speed
           control has been found to be satisfactory.

     5.6-  Flowmeter:  Capable of measuring an air flow of 1 liter  per minute.

     5.7   Aeration Tubing:  A straight glass fit having a coarse porosity.
           Tygon tubing is used for passage of the mercury vapor from  the
           sample bottle to the absorption cell and return.

     5.8   Drying Tube: 6" Z 3/4" diameter tube containing 20  g  of  magnesium
           perchlorate (see Note 2).  The apparatus is assembled as shown in
           Figure 1.

           NOTE 2:  In place of the magnesium perchlorate drying tube, a small
           reading lamp with 60V bulb may be used to prevent condensation of
           moisture Inside the cell.  The lamp is positioned to  shine  on the
           absorption cell maintaining the air temperature In  the cell about
           10°C above ambient.

6.   Reagents

     6.1   Sulfuric Acid, Cone:  Reagent grade.

           6.1.1   Sulfuric acid, 0.5 N: Dilute 14.0 ml of cone,  sulfuric  acid
                   to 1.0 liter.

     6.2   Nitric Acid, Cone:  Reagent grade of low mercury content (see Note
           3).  NOTE 3:  If a high reagent blank is obtained,  it may be  neces-
           sary eo distill the nitric acid.
                                       B-54

-------
Method 245.1  CLP-H  (cone.)
     6.3
     6.4
     6.5


     6.6


     6.7
Scannous Sulfate:  Add 25 g stannous sulface to 250 ml of 0.5
N sulfuric acid.  Th'is mixture is a suspension and should be
stirred continuously during use.  (Stannous chloride may be used
la place of stannous sulfate.)

Sodium Chloride-Hyroxylamine Sulfate Solution:  Dissolve 12 g of
sodium chloride and 12 g of hydroxylamine  sulfate in distilled
water and dilute to 100 ml.  (Hydroxylamine hydrochloride may be
used in place of hydroxylamine sulface.)

Potassium Permanganate:  5Z solution, w/v.  Dissolve 5 g of
potassium permanganate in 100 ml of distilled water.
Potassium Persulfate:  5Z solution, w/v.
persulfate In 100 ml of distilled wace.r.
                    Dissolve 5 g of potassium
Stock Mercury Solution:  Dissolve 0.1354  g  of mercuric chloride  in
75 ml of distilled water.  Add 10 ml of cone, nitric acid and
adjust the volume to 100.0 ml.  1 ml • 1  mg Hg.
                 + 'BUBBLER
       SAMPLE SOLUTION
       IN BOO BOTTLE
ABSORPTION
    CELL
                                                      SCRUBBER
                                                      CONTAINING
                                                      A MERCURY
                                                      ABSORBING
                                                      MEDIA
           Figure 1.  Apparatus  for Flaneless Mercury  Determination.
                                       3-55

-------
Method 245.1 CLP-M (cone.)


     6.8   Working Mercury Solution:  Make successive dilutions  of  the  stock
           mercury solution  to obtain a working standard  containing 0.1 ug  per
           ml.  This working standard and the dilutions of  the stock mercury
           solution should be prepared fresh daily.  Acidity of  the working
           standard should be maintained at 0.15Z nitric  acid.   This acid
           should be added to the flask as needed before  the addition of the
           aliquot.

7.   Calibration

     7.1   Transfer 0, 0.5,  1.0, 5.0 and 10.0 ml aliquots of the working
           mercury solution  containing 0 to 1.0 ug  of mercury to a  series of
           300 ml BOO bottles.  Add enough distilled water  to each  bottle to
           make a total volume of 100 ml.  Mix thoroughly and add 5 ml  of cone.
           sulfuric acid  (6.1) and 2.5 ml of concr  nitric acid (6.2) to each
           bottle. -Add 15 ml of KMn<>4 (6.5) solution to  each beetle and allow
           to stand at least 15 minutes.  Add 8 ml  of potassium  persulfate  (6.6)
           to each bottle and heat for 2 hours in a water bath maintained at
           95"C.  Cool and add 6 ml of sodium chloride-hydroxylamine sulfate
           solution (6.4) to reduce the excess permanganate.  When  the  solution
           has been decolorized wait'30 seconds, add 5 ml of the stannous
           sulfate solution  (6.3) and immediately attach  the bottle to  the
           aeration apparatus forming a closed system*  At  this  point the
           sample is allowed to stand quietly without manual agitation.  The
           circulating pump, which has previously been adjusted  to  a rate of
            1 liter per minute, is allowed  to run continuously (see  Note A).
           The absorbance will increase and reach maximum within 30 seconds.
           As soon as the recorder pen levels off,  approximately 1  minute,
           open the bypass valve and continue"the aeration  until the absorbance
            returns to its minimum value  (see Note 5).  Close the bypass valve,
            remove the stopper and frit from the BOD bottle  and continue the
            aeration.  Proceed with the standards and  construct a standard
            curve by plotting peak height versus micrograms  of mercury.

            NOTE 4:  An open  system where  the mercury  vapor  is passed through
            the absorption cell only once nay be used  instead of  the closed
            system.

            NOTE 5:  Because  of the toxic nature of  mercury  vapor precaution
            must be taken  to  avoid its  Inhalation.   Therefore, a bypass has
            been included  in  the system to  either vent the mercury vapor into
            an  exhaust hood  or  pass  the vapor  through  some absorbing media,
            such as:

                a)  equal voliimes of 0.1 M KMn04, and 10Z ^SOt
                b)  0.252 iodine  in  a  32 a KI  solution

            A specially  treated charcoal  that will adsorb  mercury vapor is also
            available  from Barnebey  and  Cheney,  E.  8th Ave.  and N. Cassidy St.,
            Columbus,  Ohio 43219,  Cat  #580-13  or  0580-22.


                                      B-56

-------
Method 245.1 CLF-H (cone.)


8.   Procedure

     S.I   Transfer 100 ml, or an aliquot diluted  to  100 ml,  containing  not
           more than  1.0 ug of mercury,  to  a  300 ml BOD bottle.  Add  5 oil  of
           eulfuric acid (6.1) and 2.5 ml of  cone, nitric acid  (6.2)  mixing
           after each addlton.  Add  15 ml of  potassium permanganate solution
           (6.5) to each sample bottle (see Note 6).   For sewage samples
           additional permanganate may be required.   Shake  and  add additional
           portions of potassium permanganate solution, if  necessary, until the
           purple color persists for at  least 15 minutes.   Add  8 ml of potassium
           persulfate (6.6) to each  bottle  and heat for 2 hours in a  water bath
           .at 95'C.

           NOTE 6:  The same amount  of KMn04  added to the samples  should be
           present in standards and  blanks.

           Cool and add 6 ml of sodium chloride-hydroxylamine sulfate (6.4) to
           reduce the excess permanganate  (see Note  7).  After  a delay of  at
           least 30 seconds add 5 ml of  stannous sulfate (6.3)  and immediately
           attach the bottle to the  aeration  apparatus.  Continue  as  described
           under Calibration.

           NOTE 7:  Add reductant in 6 mL increments  until  KMn<>4 is completely
           reduced.

9.   Calculation

     9.1   Determine  the peak height of  the unknown  from the chart and read
           the mercury  value from the standard  curve.

     9.2   Calculate  the mercury  concentration  in  the sample by the formula:

                              ug  Hg in                 1,000
                ug Hg/1 •   aliquot   X
                                               volume of aliquot in ml

      9.3   Report  mercury concentrations as follows:  Below 0.2 ug/1, 0.2U;
            between 0.2 and 10 ug/1, one decimal;  above 10 ug/1, whole numbers.
 10.   Appendix
      10.1  While the possibility of absorption from certain organic substances
            actually being present in the sample does exist, EMSU has noc encoun-
            tered such samples.  This is mentioned only to caution the analyse
            of the possibility.  A simple correction that may be used is as
            follows: If an interference has been found to be present (4.4), the
            sample should- be analyzed boch by using the regular procedure and
            again under oxidizing conditions only, that is without the reducing
            reagents.  The true mercury value can then be obtained by subtracting
            the two values.


                                       B-57

-------
     10.2   If  additional sensitivity is  required,  a 200 ml sample wich
           recorder expansion may be used provided the instrument does not
           produce undue noise.   Using a Coleman MAS-50 with a drying tube
           if  magnesium perchlorate and  a variable recorder, 2 mv was set
           to  read full scale.  With these conditions, and distilled water
           solutions of mercuric chloride at concentrations of 0.15, 0.10,
           0.05 and 0.025 ug/1 the standard deviations were +0.027, +0.0006,
           +0.01 and +0.004.  Percent recoveries at these levels were 107,
           83, 84 and~~96Z, respectively.

     10.3  Directions for the disposal of mercury-containing wastes are
           given in ASTM Standards, Part 31, "Water", p. 349, Method D3223
           (1976).
Bibliography
     1.    Kopp, J. F., Longbottom, M. C. and Lobring, L. B.  "Cold Vapor
           Method for Determining Mercury", AWWA, vol. 64, p. 20, Jan. 1972.

     2.    Annual Book of ASTM Standards, Part 31, "Water", Standard D3223-73,
           p. 343 (1976).

     3.    Standard Methods for the Examination of Water and Wastewater 14th
           Edition, p. 136 (1975).
                                       B-58

-------
                                  ATTACHMENT 5A

                                    MERCURY

              Method 245.2 CLP-M* (Automated Cold Vapor Technique)


1.   Scope and Application

     1.1   The working range is 0.2 to 20.0 ug Hg/1.

2.   Summary of Method

     2.1   The flaneless AA procedure is a physical method based on the
           absorption of radiation at 253.7 nm by mercury vapor.  The mercury
           is reduced to the elemental state and aerated from solution.  The
           mercury vapor passes through a cell*positioned in the light path of
           an atomic absorption spectrophotometer.  Absorbance'Cpeak height) is
           measured as a function of mercury concentration and recorded in the
           usual manne r.

     2.2   In addition to inorganic forms of mercury, organic mercurials may
           also be present.  These organo-mercury compounds will not respond
           to the flameless atomic absorption technique unless they are firsc
           broken down and converted to mercuric 'ions.  Potassium permanganate
           oxidizes many of these compounds, but recent studies have shown that
           a number of organic mercurials,  including  phenyl mercuric acetate
           and methyl mercuric chloride, are only partially oxidized by this
           reagent.  Potassium persulfate has been  found to give approximately
           100Z recovery when used as the oxidant with these compounds.  There-
           fore, an automated persulfate oxidation  step following the automated
           addition of the permanganate has been included to insure that organo-
           mercury  compounds, if  present, will be oxidized to  the mercuric ion
           before measurement.

 3.   Sample Handling and  Preservation

     3.1   Until more conclusive  data are obtained, samples should be preserved
           by  acidification with  nitric  acid  to  a pH  of  2  or lower immediately
           at  the time of  collection("•)  (see Exhibit  F).
 * CLP-M Modified for the contract Laboratory Program

                                       B-59

-------
Method 245.2 CLP-M (cont.)
4.   Interference (see NOTE 1)

     4.1   Some sea waters and waste-waters high in chlorides have shown  a  posi-
           tive interference, probably due Co the formation  of  free  chlorine.

     4.2   Interference from certain volatile organic materials which will  •
           absorb at this wavelength is also possible.  A preliminary run under
           oxidizing conditions, without stannous sulfate, would determine  if
           this type of interference is present.

     4.3   Formation of a heavy precipitate, in some wastewaters and effluents,
           has been reported upon addition of concentrated sulfuric  acid.   If
           this is encountered, the problem sample cannot  be analyzed by this
           method.

     4.4   Samples containing solids must be blended and then mixed while being
           sampled if total mercury values are to be reported.

           NOTE 1:  All the above interferences can be overcome by use of the
                    Manual Mercury method.

5*   Apparatus

     5.1   Technicon Auto Analyzer consisting of.:

           5,1.1  Sampler II with provision for sample mixing.

           5.1.2  Manifold.

           5.1.3  Proportioning Pump II or III.

           5.1.4  High temperature heating bath with two distillation coils
                  (Technicon Part #116-0163) in series.

     5.2   Vapor-liquid separator (Figure 1).

     5.3 '  Absorption cell, 100 mm long, 10 mm diameter with quartz windows.

     5.4   Atomic Absorption Spectrophotometer (see Note 2):  Any atomic
           absorption unit having an open sample presentation area in which
           to mount the absorption cell is suitable.  Instrument settings
           recommended by the particular manufacturer should be followed.

           NOTE 2:  Instruments designed specifically for the measurement of
                    mercury using the cold vapor technique are  commercially
                    available and may be substituted for the atomic  absorption
                    Spectrophotometer.

     5.5   Mercury Hollow Cathode Lamp:  Westinghouse VL-22847, argon filled,
           or equivalent.


                                      3-60

-------
Method 245.2 CLP-M  (cone.)


     5.6   Recorder:  Any multi-range variable speed  recorder  that  is  compat-
           ible with  the UV detection system is  suitable.

     5.7   Source of  cooling water for jacketed  mixing  coil  and  connector A-7.

     5.8   Heat lamp:  A small  reading lamp  with 60V  bulb  may  be used  to
           prevent  condensation of moisture  inside  the  cell.   The lamp is
           positioned to shine  on the absorption cell maintaining the  air
           temperature in the cell about  10'C above ambient.

6.   Reagents

     6.1   Sulfuric Acid, Cone:  Reagent  grade

           6.1.1  Sulfuric acid, 2 N: Dilute 56  ml  of cone,  sulfuric acid to
                  1 liter with  distilled  water.

           6.1.2  Sulfuric acid, 10Z:  Dilute 100 ml  cone, sulfuric acid to
                  1 liter with  distilled  water.

     6.2   Nitric acid, Cone:   Reagent grade of  low mercury  content.

           6.2.1  Nitric Acid,  0.5Z  Wash  Solution:  Dilute  5  ml of concentrated
                  nitric acid to I liter  with distilled water.'

     6.3   Stannous Sulfate:  Add-50 g stanhous  sulfate to 500 ml of 2 N
           sulfuric aeid £6.1.1).  This mixture  is  a  suspension  and should be
           stirred  continuously during use.

           MOTE 3:  Stannous chloride may be used in  place of  stannous sulfate.

     6.4   Sodium Chloride-Hydroxylamlne  Sulfate Solution:  Dissolve 30 g of
           sodium chloride and  30 g  of hydroxylamine  sulfate in  distilled
           water to 1 liter.

           NOTE 4:  Hydroxylamine hydrochloride  may be  used  in place of
           hydroxylamine  sulfate.

     6.5   Potassium  Permanganate:   0.5Z  solution,  w/v. Dissolve 5 g  of
           potassium  permanganate  in 1  liter of  distilled  water.

     6.6   Potassium  Permanganate,  0.1  N:  Dissolve 3.16 g of  potassium
           permanganate in distilled water and dilute to 1 liter.

     6.7   Potassium  Persulfate1:   0.5Z  solution, w/v.  Dissolve  5 g potassium
           persulfate in 1  liter  of  distilled water.'

     6.H    Stock Mercury Solution:   Dissolve 0.1354 g of mercuric chloride  in
            75 ml  of distilled  water.  Add 10 ml of  cone,  nitric acid  and
            adjust  the volume  to 100.0 ml. . 1.0 ml • 1.0 mg Hg.
                                        B-61

-------
Method 245.2 CLP-M (cone.)


     6.9   Working Mercury Solution:  Make successive dilutions  of  the  stock
           mercury solution  (6.8)  to  obtain  a working standard containing  0.1
           ug per ml.  This working standard and the dilutions of the stock
           mercury solution  should be prepared  fresh daily.   Acidity of the
           working standard  should be maintained at 0.15Z nitric acid.   This
           acid  should be added  to the  flask as needed  before the addition of
           the aliquot.  From this solution  prepare standards containing 0.2,
           0.5,  1.0, 2.0, 5.0, 10.0,  15.0 and 20.0 ug Hg/1.

     6.10  Air Scrubber  Solution:  Mix  equal volumes of 0.1  N potassium
           permanganate  (6.6) and  10Z sulfuric  acid (6.1.2).

 7.   Procedure

     7.1   Set up manifold as shown in  Figure 2.

     7.2   Feeding  all the reagents through  the system  with  acid wash solution
           (6.2.1)  through the sample line,  adjust heating bath  to  105°C.

     7.3   Turn  on  atomic absorption  spectrophotometer, adjust instrument
           settings  as recommended by the manufacturer, align absorption' cell
           in light  path for maximum  transmlttance  and  place heat lamp  directly
           over  absorption cell.

     7.4   Arrange working mercury standards from 0.2 to 20.0 ug Hg/1 in sampler
           and start sampling.  Complete  loading of  sample  tray with unknown
           samples.

      7.5   Prepare  standard  curve by  plotting  peak  height of processed  standards
           against  concentration values.  Determine concentration  of samples  by
           comparing sample  peak height with standard curve.

            NOTE 5:   Because  of the toxic nature of mercury vapor,  precaution
           must be  taken to  avoid its Inhalation.   Venting the mercury  vapor
            into an exhaust hood or passing the vapor through some  absorbing
            media such as:

            a)  equal volumes of 0.1  N KMn04(6.6)  and 10Z H2S04 (6.1.2).
            b)  0.25Z iodine in a  32 KI solution,  is recommended.

            A specially treated charcoal that will adsorb mercury vapor is  also
            availale from Barnebey and Cheney, E.  8th Ave. and North Cassidy St.,
            Columbus, Ohio 43219,  Cat, #580-13 or #580-22.
      7.6   After the analysis if complete put all lines except the H^SOt line
            in distilled water to wash out system.  After flushing, was out the
            H2S04 line.  Also flush the coils in the high temperature heating bach
            by pumping stannous sulfate (6.3) through the sample lines followed
            by distilled water. This will prevent build-up of oxides of manganese.
                                       B-62

-------
Method 245.2 CLP-M (cont.)
Bibliography

1.   Wallace R. A., Fulkerson, W., Shults, W. D., and Lyon, W.  S.,  "Mercury
     in  the Environment-The Human Element",  Oak  Ridge National  Laboratory,
     ORNL/NSF-EP-1 p.  31,  (January,  1971).

2.   Hatch, W. R. and  Ott, W.  L., "Determination of  Sub-Microgram Quantities
     of  Mercury by Atomic Absorption Specrophotometry".   Anal.  Chem.  40,  2085
     (1968).

3.   Brandenberger, H.  and Bader, H.,  "The Determination of Nanogram Levels
     of  Mercury in Solution by a Flameless Atomic Absorption  Technique",
     Atomic Absorption Newsletter £, 101  (1967).

4.   Brandenberger, h.  and Bader, H.,  "The Determination of Mercury by
     Flameless Atomic  Absorption II, A Static Vapor  Method",  Atomic
     Absorption Newsletter J.,53 (1968).

5.   Goulden, P.  D. and Afghan,  B. K.  "An Automated  Method for  Determining
     Mercury  in Water", Technlcon, adv. in Auto. Analy.   .2,  p.  317 (1970).

6.   "Interim Methods  for  the Sampling and Analysis  of  Priority Pollutants  in
     Sediments and Fish Tissue," USEPA, Environmental Monitoring and Support
     Laboratory,  Cincinnatin, Ohio,  august  1977,- revised Occtober 1980.

 7.   Op. clt. (#1), Methods  245.1 or 245.2.
                                        B-63

-------
Method 245.2 CLP-M (cone.)
      AIR AND
      SOLUTION
      M
                                                  AIR
                                                  OUT
                                   7/23 T
0.7 cm 10
                      0.4 em ID
14 cm
                                                      SOLUTION
                                                       OUT
                     Figure 1.  Vapor liquid separator.
                                  B-64

-------
               AA

          SPECTROPNOTOMETER
      Dl



MflflftflOO
         0000
                               lUnn Ilia

                                  Long
                                  251. 7nn
                                                                     20/hr
                                          RECORDER
                                V«nt
                           Waste
                                    €:
                                       S-3
09
I
5
          HEATING BATH
                     I PSI
                                                    i;
                                                           r.
                                                             l/.in
                                                             3.90
                                                      2.00 |'DO
                              3
      CO MM03 Uash
                                                                                        g-
                                                                                        p.
                                                             3.90
                                                      SFi
                                                             ^.90
                                                             3.90
                                                      JJ2.
                                                             3.90 co
                                                             i.20
                            1
                                                                 Air
               AIR

             SCRUBBER
                                                              10Z SnS0
                                                                           4
                                                                    3Z NaCl

                                                                       4
                                                             J.50
                                                             1.20
                                                                K?S]0|
                                                             2.76
                                                                    (Con)

                                                             1.90
                                                             l.l'l
:?•
                                                               Sanplc
                                                                                        n
                                                                                        o
                                                                                        a
                                                             2JDO
                                                               Air
                                                             1.30
                                                             .51  KHn06
                                                Proportioning Pump III



                              Figure 2. Mercury .Manifold AA-1.
                                                              *  Acid Flem iCRUBBEft

                                                              ** Teflon

                                                             ••* Claa«

-------
                                  ATTACHMENT 6

                             MERCURY (in Sediments)

               Method 245.5 CLP-M* (Manual Cold Vapor Technique)


1.   Scope and Application

     1.1   This procedure measures total mercury (organic and inorganic) in
           soils, sediments, bottom deposits and sludge type materials.

     1.2   The range ofrthe method is 0.2 to 5 ug/g.  The range may be extended
           above or below the normal range by increasing or decreasing sample
           size or through instrument and recorder control.

2.   Summary of Method'

     2.1   A weighed portion of the sample is acid digested for 2 minutes at
           95'C, followed by oxidation with potassium permanganate and potassium
           persulfate.  Mercury in the digested sample is then measured by the
           conventional cold vapor technique.

     2.2   An alternate, digestion involving the use of an autoclave is describe
           in (a.2).

3.   Sample Handling and Preservation

     3.1   Because of the extreme sensitivity of the analytical procedure and
           the omnipresence of mercury, care- must be taken to avoid extraneous
           contamination.  Sampling devices and sample containers should be
           .ascertained to be free of mercury; the sample should not be exposed
           to any condition in the laboratory that may result in contact or
           air-borne mercury contamination.

     3.2   Refrigerate solid sanples upon receipt.

     3.3   The sample should be analyzed without drying.  A separate  I solids
           determination is required.   (Exhibit D, Attachment 9).

4.   Interferences

     4.1   The same  types of interferences that may occur in water samples are
           also  possible with sediments,  i.e., .sulfides, high copper, high
           chlorides, etc.
 *CLP-M modified  for  the  Contract  Laboratory  Program

                                     B-66

-------
Method 245.5 CLP-M  (cone.)


     4.2   Volatile materials which absorb at  253.7 nm will  cause  a  positive
           interference.  In order to  remove any  interfering volatile  materials,
           purge the dead air space in the BOD bottle before the addition  of
           stannous sulfate.

     4.3   Sample containing high concentrations  of oxidizable organic materials,
           as evidenced by high chemical oxygen demand values, may not be  com-
           pletely oxidized by this procedure.  When this occurs,  the  recovery
           of organic mercury will be  low.  The problem  can  be eliminated  by
           reducing the weight of the original sample or by  Increasing the
           amount of potassium persulfate (and consequently  stannous chloride)
           used in the digestion.

5.   Apparatus

     5.1   Atomic Absorption Speetrophotometer (see Note 1):  Any  atomic absorp-
           tion unit having an open sample presentation area In which  to mount
           the absorption cell is suitable.  Instrument settings recommended by
           the particular manufacturer should be  followed.

           NOTE 1:  Instruments designed specifically for the measurement of
           mercury using the cold vapor technique are commercially available
           and may be substituted for the atomic absorption  spectrophotometer.

     5.2   Mercury Hollow Cathode Lamp:  Westlnghouse WL-22847, argon  filled,
           or equivalent.

     5.3   'Recorder:  Any multi-range variable speed recorder that is  compatible
           with the UV detection system is suitable.

     5.4   Absorption Cell:   Standard spectrophotometer cells 10 cm long, having
           quartz end windows may be used.  Suitable cells many be constructed
           from pexlglass tubing, 1" O.D. X 4-1/2".  The ends are ground perpen-
           dicular to the longitudinal axis -and quartz windows U" diameter X
           1/16" thickness) are cemented in place.  Gas inlet and outlet ports
           (also of plexiglass but 1/4" 0.0.) are attached approximately 1/2"
           from each end.  The cell is strapped to a burner  for support and
           aligned in the light beam to give the maximum transmittance.

           NOTE 2:  Two 2" X 2" cards with one inch diameter holes may be
           placed over each end of the cell to assist in positioning the cell
           for maximum transmittance.

     5.5   Air Pump:  Any peristaltic pump capable of delivering 1 liter or air
           per minute may be used.  A Masterflex pump with electronic speed
           control has been found to be satlsfatory.   (Regulated compressed air
           can be used in an open one-pass system'.}

     5.6   Flowmeter:   Capable of measuring an air flow of 1  liter per  minute.
                                       B-67

-------
Method 245.5 CLP-M (cont.)


     5.7   Aeration Tubing:  Tygon tubing is used for passage  of  the mercury
           vapor from the sample bottle to  the absorption  cell and return.
           Straight glass tubing terminating in a coarse porous frit is  used
           for sparging air into the sample*

     5.8   Drying Tube:  6" X 3/4" diameter tube containing  20 g  of magnesium
           perchlorate (see Note 3).  The apparatus  is assembled  as shown in
           the accompanying diagram.

           NOTE 3:  In place of the magnesium perchlorate  during  tube, a small
           reading lamp with 60V bulb may be used to prevent condensation of
           moisture inside the cell.  The lamp Is positioned to shine on the
           absorption cell maintaining the  air temperature in  the cell about
           1U*C above ambient.

6.   Reagents

     6.1   Sulfuric acid, cone.:  Reagent grade of low mercury content.

     6.2   Nitric acid, cone.:  Reagent grade of low mercury content.

     6.3   Stannous Sulfate:  Add 25 g stanaous sulfate to 250 ml of 0.5 N
           •ulfuric acid (6.2).  This mixture is a suspension  and should be
           stirred continuously during use.

     6.4   Sodium Chloride-Hydroxylamine Sulfate Solution:   Dissolve 12  g of
           sodium chloride and 12 g of hydroxylamine sulfate in distilled water
           and dilute to 100 ml.

           NOTE 4:  A 102 solution of stannous chloride may  be substituted
           for (6.3) and hydroxylamine hydrochloride may be  used  in place of
           hydroxylamine sulfate In (6.4).

     6.5   Potassium Permanganate:  52 solution, v/v.  Dissolve 5 g of potassium
           permanganate in 100 ml of distilled water.

     6.6   Potassium Persulfate:  52 solution, w/v.  Dissolve  5 g of pocassium
           persulfate In 100 ml of distilled water.

     6.7   Stock Mercury Solution:  Dissolve 0.1354  g of mercuric chloride in
           75 ml of distilled water.  Add ml of cone, nitric acid and adjust
           the volume to 100.0 ml.  1.0 - 1.0 mg Kg.

     6.8   Working Mercury Solution:  Make  successive dilutions of the stock
           mercury solution (6.7) to obtain a working standard containing 0.1
           ug/ml.  This working* standard and the dilution  of the  stock mercury
           solutions should be prepared fresh dally. Acidity  of  the working
           standard should be maintained at 0.152 nitric acid. This acid
           should be added to the flask as  needed before the addition of che
           aliquot.
                                      B-68

-------
Method 245.5 CLP-M (coat.)
7.   Calibration

     7.1   Transfer 0, 0.5, 1.0, 5.0 and 10 ml aliquots of  Che working  mercury
           solutions  (6.8) containing 0 to 1.0 ug  of  mercury  to  a series  of  300
           ml BOD bottles.  Add enough distilled water to each bottle to  make a
           total volume of 10 ml.  Add 5 ml of cone.  ^SOfc  (6.1) and 2.5  ml  of
           cone. H1K>3  (6.2) and heat 2 minutes in  a water bath at 95°C.  Allow
           the sample  to cool and add 50 ml distilled water,  15  ml of KMn04
           solution (6.5) and 8 ml of potassium persulfate  solution (6.6) to
           each bottle and return to the water bath for 30  minutes. Cool and
           add 6 ml of sodium chloride-hydroxylamine  sulfate  solution  (6.4)  to
           reduce the  excess permanganate.  Add 50 ml of distilled water.
           Treating each bottle individually, add  5 ml of stannous sulfate
           solution (6.3) and Immediately attach the  bottle to the aeration
           apparatus.  At this point the sample is allowed-Co stand quietly
           without manual agitation.  The circulating pump, which has previously
           been adjusted to a rate of  1 liter per  minute, is  allowed to run
           continuously.  The absorbance, as exhibited either on the spectro-
           photometer or the recorder, will Increase  and reach maximum  within
           30 seconds. As soon as the recorder pen levels  off,  approximately  1
           minute, open the bypass valve and continue Che aeration until  the
           abscrbance returns to its minimum value (see Note  5).  Close the
           bypass valve, remove  the  fritted tubing from  the BOD  bottle  and
           continue the aeration.  Proceed with the standards and construct  a
           standard curve by plotting  peak height  versus micrograms of  mercury.

           NOTE 5:  Because of the toxic nature of mercury  vapor precaution
           must be  taken to avoid  its  inhalation.   Therefore, a  bypass  has been
           included in the system  to either vent Che  mercury  vapor into an
           exhaust hood or pass  the  vapor  through  some absorbing media, such as:

           a)  equal  volumes of  0.1  N  KMn04 and  10Z H2S04
           b)  0.25Z  iodine  in  a 3Z  KI solution

           A specially treated  charcoal  that will  absorb mercury vapor  is also
           available  from  Barnebey and Cheney,  E.  8th Avenue  and N. Cassidy
           Sreet, Columbus, Ohio 43219

 b.    Procedure

      8.1  Weigh  a  representative  0.2  g  portion  of wet sample and place in the
            bottom of  a BOD bottle.   Add 5  mL  of  sulfurlc acid (6.1) and 2.5  mL
           of concentrated nitric  acid (6.2) mixing after  each addition.   Heat
            two  minutes in  a  water bath at  95*C.   Cool,  add  50 ml distilled
           water,  15  mL  potassium permanganate  solution  (6.5) and 8 mL of
            potassium persulfate solution (6.6)  to  each sample bottle.   Mix
            thoroughly and  place  in the water  bath  for 3U minutes at 95°C.  Cool
            and  add 6  mL  of  sodium chloride-hydroxylamine sulfate (6.4)  to reduce
            the  excess permanganate.   Add 55  mL  of  distilled water.  Treating
            each bottle individually, add 5 mL of  stannous  sulfate  (6.3) and
            immediately attach  the  bottle to  the  aeration apparatus.  Continue
            as described  under (7.1)..
                                     8-69

-------
Method 245.5 CLP-M (cent.)


     8.2   An alternate digestion procedure employing an autoclave  may  also be
           used.  In this method 5 ml of cone* ^SQ^ and 2 ml  of  cone.   HN03
           are added Co the 0.2 g of sample.  5 ml of saturated KMnO^ solution
           and 8 ml of potassium persulfate solution are added and  the  bottle is
           covered with a piece of aluminum foil.  The sample is  autoclaved at
           121*C and 15 Ibs. for 15 minutes.  Cool, make up  to a  volume of 100
           ml with distilled water and add 6 ml of sodium chloride-hydroxylamine
           sulfate solution (6.4) to reduce the excess permanganate.  Purge the
           dead air space and continue as described under (7.1).

9.   Calculations

     9.1   Measure the peak height of the unknown from the chart  and read the
           mercury value from the standard curve.

     9.2   Calculate the mercury concentration in the sample by the formula:

                       ug Hg in the aliquot
           ug Bg/g -   wt of the aliquot in gms (based upon dry wt of the sample)

     9.3   Report mercury concencrations as follows:  Below 0.1 ug/gm,  0.1U;
           between 0.1 and 1 ug/gm, to the nearest 0.01 ug; between 1 and 10
           ug/gm, to nearest 0.1 ug; above 10 ug/gm, to nearest ug.

Bibliography

     1.    Bishop, J. N., "Mercury in Sediments'", Ontario Water Resources
           Conm., Toronto, Ontario, Canada, 1971.

     2.    Salma, M., private communication,  EPA Cal/Nev. Basin Office,
           Almeda, California.

     3.    "Interim Methods for the Sampling and Analysis of Priority Pollutants
           in Sediments and Fish Tissue," USEPA Environmental Monitoring and
           Support Laboratory, Cincinnati,  Ohio, August 1977, Revised October
           1980.

     4.    Op.  eit. («), Methods 245.1 or 245.2.
                                      B-70

-------
                  EZHIBIT E




QUALITY ASSURANCE/QUALITY CONTROL REQUIREMENTS
                      B-7I

-------
                        GENERAL QA/QC CONSIDERATIONS
     Standard laboratory practices for laboratory cleanliness as applied
to glassware and apparatus oust be adhered to.  Laboratory practices with
regard to reagents, solvents, and gases should also be adhered to.  For
additional guidelines regarding these general laboratory procedures, please
see Sections 4 and 5 of the Handbook for Analytical Quality Control in
Water and Wastevater Laboratories EPA-600/4-79-019, USEPA Environmental
Monitoring and Support Laboratory, Cincinnati, Ohio, March 1979.
                                    B-72

-------
                         QUALITY ASSURANCE REQUIREMENTS


The quality assurance/quality control (QA/QC.) procedures defined below must
be used by the Contractor when performing the methods  specified in  Exhibit
D.  When additional QA/QC procedures are specified in  the methods in Exhibit
D, the Contractor must also follow these procedures.   The cost of performing
all QA/QC procedures specified in this Statement of Work is included in  the
price of performing the bid lot.

The purpose of this document is to provide a uniform set of procedures for
the analysis of inorganic constituents of samples, documentation of methods
and their performance during a survey, and verification of the sample data
generated.  The program will also assist laboratory personnel in recalling
and defending their actions under cross examination if required to  present
court testimony in enforcement ease litigation.

The prime function of the QA/QC program outlined here  is the definition  of
procedures for the evaluation and documentation of sampling and analytical
methodologies and the reduction and reporting of data. The objective is to
provide a uniform basis for sample collection and handling, instrument and
methods maintenance, performance evaluation, and analytical data gathering
and reporting.  Although it is impossible to address all analytical situations
In one document, the approach taken here is to define  minimum requirements
for all major steps relevant to any inorganic analysis.  In many instances
where methodologies are available, specific quality control procedures are
Incorporated into the method documentation.  Ideally,  samples involved in
enforcement actions are analyzed only after the methods have met the minimum
performance and documentation requirements described in this document.

The Contractor must participate In the Laboratory Audit and Intercomparlson
Study Program run by  EPA, EMSL-Las Vegas.  The Contractor  can expect  to
analyze two samples per three-month contract period for this program.

The Contractor must perform and report to SMO and EMSL/LV  on Form XI  quarterly
verification of  instrument  detection  limits  (IDL)  by  the method  specified in
Exhibit E, by type and model for each instrument used  on this contract.   All
 the  IDLs  must meet  the  CXDLs specified  in Exhibit  C.   For  ICP methods,  the
Contractor must  also  report linearity range verification,  all interelement
 correction factors, wavelengths used  and integration  times on a  quarterly
 basis on  Forms XII  and  XIII.

 For  QA/QC procedures,  two  different  types of  "samples" are specified.  A
 "sample received"  is  the field or PE  sample  each of which  has  an EPA number.
An "analytical  sample"  is  each "analysis"  performed  (i.e.  each  cup, tube or
 container in autosamplers,  etc),.   A "frequency of  102" means  once  every 10
 analytical  samples.

 The  Contractor must report all QC data in the exact  format shown in Forms
 II - XIII (See Exhibit B).
                                       B-73

-------
This section outlines the minimum QA/QC operations necessary to satisfy  the
analytical requirements of the contract.  The following (JA/QC operations must
be performed as stated in this exhibit:

     1.  Initial Calibration and Calibration Verification

     2.  Continuing Calibration Verification

     3.  Preparation Blank Analysis

     4.  Interference Check Sample Analysis

     5.  1C? Serial Dilution Analysis

     6.  Matrix Spike Analysis

     7.  Duplicate Sample Analysis

     8.  Furnace AA QC Analysis (Method of Standard Additions may be required
         under certain conditions).

     9. Laboratory Quality Control Sample Analysis

1.   Initial Calibration and Calibration Verification

     Guidelines for instrumental calibration are given in EPA 600/4-79-020
     and/or Exhibit D.  Instruments must be calibrated daily and each time
     the instrument is set up*

     Instrument Calibration

     For atomic absorption systems, calibration standards are prepared by
     diluting the stock mecal solutions at the time of analysis.  Low calibra-
     tion standards must be prepared fresh each time an analysis is to be made
     and discarded after use.  Prepare'a blank and at least three calibration
     standards in graduated amounts in the appropriate range*  One atomic
     absorption calibration standard must be at the CSDL except for mercury.
     The calibration standards must be prepared using the same type of acid- or
     combination of acids and at the same concentration as will result in the
     samples following sample preparation.  Beginning with the blank and working
     toward the highest standard, aspirate or Inject the solutions and record
     the readings.  Calibration standards for furnace- procedures should  be
     prepared as described in the individual methods for that metal.

     For cyanide and mercury, follow the calibration procedures outlined  in
     Exhibit D.  One cyanide calibration standard must be- at  the CSDL.   For 1C?
     •y8cams, calibrate the instrument according to instrument manufacturer's
     recommended procedures.

     Initial Calibration Verification  Standards

     After  the  ICP, AA and cyanide  systems have been calibrated,  the  accuracy
     of the initial calibration shall  be verified and documented  for  every
     analyce  by the analysis of EPA Initial  Calibration Verification  Solution.
                                       8-74

-------
The initial calibration verification standard must be run at each wave-
length used for analysis.

If the Initial Calibration Verification Solution is not available from
EPA, or where a certified solution of an analyte is not available from any
source, analyses shall be conducted on an independent standard at a concen-
tration other than that used for calibration, but within the calibration
range*  An independent standard is defined as a standard composed of the
analytes from a different source than those used in the standards for the
initial calibration.  For ICP, the Initial Calibration Verification Solu-
tion must be run at each wavelength used in the analysis of the sample.
When measurements exceed the control limits of Table 1, the analysis must
be terminated, the problem corrected, the instrument recalibrated, and the
calibration reverified.

The values for the Initial and subsequent continuing calibration veriflea-
ions shall be recorded on Form II (see Exhibit B) for ICP, AA, and cyanide
analyses, as indicated.

Instrument Detection Limit Determination

Before any field samples are analyzed under this contract, the•instrumental
detection limits (in ug/L) must be determined within 30 days of the start
of the analyses and at least quarterly (every 3 months), and must meet the
levels specified in Exhibit C.  The instrumental detection limits (in ug/L)
•hall be determined by multiplying by 3, the average of the standard devia-.
tions obtained on three nonconsecutive days from the analysis of a standard
solution (each analyte in reagent water) at a concentration 3-5 times IDL,
with 7 consecutive measurements per day*

For each Case, IDLs must be reported on QC Report Fora VII.  If multiple
AA Instruments are used for the analysis of an element within a Case, the
highest IDL for the AAs must be reported for that Case.  Use the same
reporting procedure for multiple ICPs.

Linear Range Analysis

To verify linearity near the CRDL for ICP analysis, the contractor must
analyze an ICP standard at two times the CRDL at the beginning and end of
each sample analysis run, or a minimum of twice per 8 hour working shift,
whichever Is more frequent.  This standard must be run for all elements
analyzed by ICP except Al, Be,- Ca, Fe, Mg, Ha and K.

For ICP analysis, a linear range verification check standard must be
analyzed and reported quarterly for each element on Form XII.  The standard
must be analyzed during a routine analytical run performed under this con-
tract.  The analytically determined concentration of this standard must be
wiehln +_ 52 of the true value.  This concentration is the upper llmlc of
the ICP~iinear range beyond which results cannot be reported under this
contract.
                                 B-75

-------
     Calibration Blank

     A calibration blank is analyzed each time Che instrument is calibrated, at
     the beginning and the end of the run, and at a frequency of 10Z during the
     run.  The results for the calibration blank solution shall be recorded on
     Fora III for ICP, AA and cyanide analyses, as indicated.  Blanks are to be
     reported down to the instrument detection limit.  If this blank result is
     greater than the CRDL (Exhibit C), terminate analysis, correct the problem
     and recalibrate.

2.   Continuing Calibration Verification

     To assure calibration accuracy during each analysis run, one of the follow
     ing standards is to be used for continuing calibration verification and must
     be analyzed for each analyte at a frequency of 10Z or every 2  hours during
     an analysis run, whichever is more frequent.  The standard must also be
     analyzed for each analyte at the beginning of the run and after the last
     analytical sample.  The analyte concentrations in the continuing calibration
     standard must be one'of the following solutions at or near the mid-range
     levels of the calibration curve:

     1.   EPA Solutions

     2.   NBS SUM 1643a

     3.   A contractor-prepared standard solution
           TABLE 1.  INITIAL AND CONTINUING CALIBRATION VERIFICATION
                     CONTROL LIMITS FOR INORGANIC ANALYSES
                                                    Z of True Value  (EPA Set)
 Analytical Method       Inorganic Species          Low Limit      High Limit
ICP/AA
Cold Vapor AA
Other
Metals
Mercury
Cyanide
90
80
85
110
120
115
The same continuing calibration standard must be used throughout  the  analysis
runs for a case of samples  received.

If the deviation of the  continuing  calibration verification  is  greater  than
the Control Limits specified  in Table  1, the instrument  must be recalibrated
and the preceding  10  samples  reanalyzed for the analyses  affected.   Information
regarding the  continuing verification  of calibration shall be recorded  on Fora
II (see Exhibit B) for  ICP, AA and  cyanide as indicated.

                                       B-76

-------
3.   Preparation Blank Analysis

     At least one preparation blank (or reagent blank), consisting  of deionized
     distilled water processed through each sample preparation procedure  (i.e.
     water, solids) performed for each Case, must be prepared and analyzed with
     every 20 samples received, or with each batch* of samples digested whichever
     is more frequent (see Exhibit D).  The first 20 samples of a Case are to be
     assigned to preparation blank one,'and the second 20 samples to preparation
     blank two, etc. (see Form III).  Each data package must contain'the  results of
     all the preparation blank analyses associated with the samples in that Case.

     This blank is to be reported for each Case and used in all analyses  to
     ascertain whether sample concentrations reflect contamination  in the
     following manner:

     1.   If the concentration of the blank is less than or equal to the  con-
          tract required detection level (Exhibit C), no correction of sample
          results is performed.

     2.   If the concentration of the blank is above the contract required de-
          tection level:   For any group of samples associated with a particular
          blank, the concentration of the sample with the least concentrated
          analyte oust be 10X the blank concentration, or all samples associated
          with the blank and less than 10 times the blank concentration must be
          redigested and reanalyzed, with the exception of an Identified  aqueous-
          soil field blank.  The sample value is not to be corrected for  the
          blank value.

          The values for the preparation blank shall be recorded in ug/L  on Fora
          III (see Exhibit B) for ICP, AA, and cyanide analyses, as indicated.

4.   ICP Interference Cheek Sample Analysis

     To verify inter-element and background correction factors the Contractor
     oust analyze and report the results for an 1C? Interference Check Sample at
     the beginning and end of each sample analysis run (or a minimum of twice per
     8 hour working shift, whichever is most frequent).  The ICP Interference
     Cheek Sample must be obtained from EPA (EMSL/LV) if available.  Results for
     the check sample analysis during the analytical runs must fall within Che
     control Halt of ±202 of the true value for the analytes included in the
     Interference Check Sample.  If not, terminate the analysis, correct  the
     problem, recalibrate, reverlfy the calibration, and reanalyze the samples.
     If the ICP Interference Check Sample is not available from EPA, an independ-
     ent ICP Cheek Sample must be prepared with interferent and analyte concen-
     trations at the levels specified in Table 2 (Exhibit E).  The mean value and
     standard deviation must be 'established by initially analyzing the check sam-
     ples at lease 5 times repetitively for each parameter on Form  IV.  Resulcs
     must fall within the control limit of +20Z of the established mean value.

     Results of all Interference Check Sample analyses for all ICP parameters
     must be recorded on Form IV (see Exhibit B).
*A group of samples prepared ac  the same time.


                                      B-77

-------
5.   1C? Serial Dilution Analysis

     Prior to reporting concentation data for the analyte elements, the Con-
     tractor must analyze and report the results of Che ICP Serial Dilution
     Analysis.  The ICP Serial Dilution  Analysis must be performed on each
     group of samples of a similar matrix type  (i.e. water, soil) and concen-
     tration (i.e. low, medium) for each Case of samples, or for each 20 samples
     received, whichever is more frequent.  Samples, identified  as field blanks
     cannot be used for serial dilution analysis.  If the analyte concentration
     Is sufficiently high (minimally a factor of 10 above the instrumental
     detection limit after dilution), an analysis of a 1:4 dilution must agree
     within 10 percent of the original determination.  If the dilution analysis
     Is not within 10Z, a chemical or physical  Interference effect should be
     suspected, and the data must be flagged with an "E."  Serial dilution
     results must be reported on QC Report Form IX.

6.   Spiked Sample Analysis

     The spiked sample analysis is designed to  provide information about the
     effect of che sample matrix on the digestion and measurement methodology.
     The spike is added before the digestion and prior to any distillation
     steps (I.e., CH~).  At lease one spiked sample analysis must be performed
     on each group of samples of a similar matrix type (i.e. water, soil) and
     concentration (i.e. low, medium) for each  Case of samples  or for each 20
     samples received, whichever is more frequent.*  Samples identified as
     field blanks cannot be used for spiked sample analysis.  The analyte spike
     must be added in Che amount given in Table 3 (Exhibit E) for each element
     analyzed.  If. two analytical methods are used to. obtain the reported
     values for the same element for a Case of  samples (i.e. 1C?, GFAA), spike
     samples must be run by each method used.   If the spike recovery is not
     within the limits of 75-125Z, the data of  all samples received associated
     with that spiked sample must be flagged with the letter "N".  An excep-
     tion to this rule is granted in situations where the sample concentration
     exceeds the  spike concentration by a factor of four or more.  In such a
     case, the spike recovery should not be. considered and the  data shall be
     reported unflagged even if the percent recovery does not meet the 75-125Z
     recovery criteria.  In the instance where  there is more than one spiked
     sample per matrix per Case, if one spike sample recovery is not within
     contract criteria, flag all the samples of the same matrix in the Case.
     Individual component percenc recoveries (ZR) are calculated as follows:

                  ZRecovery  -   (SSR-SR)    x   100
                                   SA

     Where   SSR   -   Spiked  Sample Result
              SR   •   Sample Result
              SA   -   Spike Added

     When  sample  concentration is less  than the instrument detection  limit,  use
      SR •  0  for  purposes  of  calculating ZRecovery.  The  spiked  sample  results
     must  be  reported  on  Form V  for  ICP, AA and cyanide  analyses,  as  indicated.

 *EPA may require  additional  spike sample analysis,  upon  request,  for which  the
  contractor  will  be paid.


                                      B-78

-------
TABLE 2.  1NTERFERENT AND ANALITE ELEMENTAL CONCENTRATIONS USED FOR  1C?
          INTERFERENCE CHECK SAMPLE
Analytes
Ba
Be
Cd
Co
Cr
Cu
Mn
Ni
Fb
V
Zn
(mg/L)
0.5
0.5
1.0
0.5
0.5
0.5
0.5
1.0
1.0
0.5
1.0
Interferents
AI
Ca
Fe
Hg







(mg/L)
500
500
200
500







              TABLE 3.  SPIRING LEVELS1 FOR SPIKED SAMPLE ANALYSIS
Element
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Mercury
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc
Cyanide
For ICP/AA
(ug/L)
Water
2,000
500

2,000
50
50
*
200
500
250
1,000
500
*
200

400
*

50
*

500
200

Sediment-1
*
500

2,000
50
50
*
200
500
250
*
500
*
500

500
*

50
*

500
500

For
Water

100
20


5





20





10


50



Furnace AA
(ug/L)
Sediment1
.
100
40


5





50





10


50



Other
(ug/L)














1








100
NOTE:  Elements without spike levels and not designated with an asterisk, should
be spiked at appropriate levels t
     levels shown indicate concentrations in the final dlgestate of the spiked
 sample (200 mL FV)
 No spike required.
                                      B-79

-------
    Duplicate Sample Analysis

    Ac least one duplicate sample must be analyzed  from each group of  samples
    of a similar matrix  type  (i.e. water, soil)  and concentration (i.e.   low,
    medium) for each Case of  samples or for  each 20 samples  received,  whichever
    is more frequent.*   Samples  identified as  field blanks cannot be used for
    duplicate sample analysis.   If two analytical methods  are used to  obtain the
    reported values for  the same element for a Case of samples (i.e. ICP, GFAA),
    duplicate samples must be run by each method used.  The  relative percent
    differences'*'  (RPD) for each  component are  calculated as  follows:
                     RPD  -       "      ,  100
     Where RPD  •  Relative Percent Difference
           D}   •  First Sample Value
           D£   •  Second Sample Value (duplicate)

     The results of the duplicate sample analysis must be reported on Form VI.
     (See, Exhibit B). A control limit of ± 20Z for  RPD shall be used for sample
     values greater than 5 times the contract required detection level (CRDL).
     A control limit of + the CRDL shall be used for sample values less than
     5 times the CRDL (Exhibit C), and this control limit (+CRDL) should be
     entered in the "Control Limit" column on Form VI.  If one result is above
     the 5 x CRDL level and the other is below, use the + CRDL criteria.  If
     either sample value is less than the CRDL, the RPD is not calculated and
     is indicated as "NC" on Fora VI.

     If the duplicate 'sample results are outside the control limits, flag all
     the data for samples received associated with that duplicate sample with
     an "*' on Forms I and V.  In the instance where there is more than one
     duplicate sample per Case, if one duplicate result is not within contract
     criteria, flag all the samples of the same matrix in the Case.  The percent
     difference data will be used by EPA to evaluate the long-term precision of
     the methods for each parameter.  Specific control limits for each element
     will be added to Form VI at a later date based on these precision results.

8.   Furnace Atomic Absorption (AA) QC Analysis

     Because of the nature of the Furnace AA technique, the special procedures
     summarized in Figure 1 will be required for quantltation.  (These procedures
     are not meant to replace those included in Exhibit D of this document, but
     will supplement  the guidance provided therein.)
*EPA may require additional duplicates be analyzed, on request, for
 which the contractor will be paid.

^Relative percent difference is equivalent to relative range of duplicates (RR).


                                     B-80

-------
                                     Figure 1.

                    FURNACE ATOMIC ABSORPTION ANALYSIS SCHEME
 PREPARE AND ANALYZE SAMPLE
 AND ONE SPIKE (2 X C.R.D.L.)
 (double injections required)
 ANALYSIS WITHIN CALIBRATION RANGE
                                                NO
                                                                 DILUTE  SAMPLE
               YES
 RECOVERY OF SPIKE >40%
               YES
                                              if NO,  repeac  only  once
                           if StillwI  BTJIC  DATA UTTH AN  "E"
               NO
 SAMPLE ABSORBANCE >50Z of  SPIKE ABSORBANCE*
                                                 NO
               YES
  SPIKE  RECOVERY  <85Z  OR >115Z
                                                 NO
             1
YES
  QUANTITATE BY MSA WITH 3 SPIKES
  AT 50,  100 AND 150Z OF SAMPLE
         ABSORBANCE
  (only single injections required)
  CORRELATION COEFFICIENT >0.995
                                    if NO,
                                    repeac only once
                YES
       FLAG DATA WITH "a
                               if still NO
                                                 REPORT SAMPLE AS
                                                 
-------
    1.   All furnace analyses, except  during  Full Methods  of Standard Addition
         (MSA), will require duplicate injections for which the average absorb-
         anee or  "concentration" will  be  reported.   All analyses must fall
         within the calibration range.  The raw data package oust contain
         both absorbance or "concentration" values,  the average value and che
         relative  standard deviation  (RSD) or'coefficient  of variation (CV).
         For concentrations greater than  CRDL,' che  duplicate injection readings
         oust agree within 202 RSD or  CV, or  che sample must be rerun once.
         If the readings are still out, flag  the value with an "M" on Form I.

    2.   All furnace analyses for each sample will  require at least a single
         analytical spike to determine if the MSA will be  required for quanci-
         tation.   Analytical spikes  are  not  required on the predlgest spike
         samples.  The spike* will be  required to be at a  concentration (in
         the sample) twice the contract required detection limit (CRDL).  The
         percent  (ZR) of the spike, calculated by the sane formula as Spiked
         Sample analyses (Exhibit E),  will then determine  how the sample will
         be quantitated as follows:

         a)  If the spike recovery is  less than 40Z, the saaple must be diluted
             and  rerun with another spike.  Dilute  che sample by a factor of  5
             to 10 and rerun.  This seep  oust only  be performed once.  If after
             the  dilution the spike recovery  is still <40Z, report data and
             flag with an "E" to indicate interference problems.

         b)  If the spike recovery is-greater than  40Z and the sample absorbance
             or concentration is <50Z  of  the  spike4", report the sample as less
             than the CRDL or less than the CRDL times the dilution factor if
             the  sample was diluted.

         c)  If the sample absorbance  or  concentration is  >50Z of the spike*
             and  the spike  recovery is between 85Z  and 115Z, the sample should
             be quantitated directly  from the calibration  curve.

         d)  If the sample absorbance or  concentration is  >50Z of the spike*
             and  the spike  recovery is less  than 85Z or greater than 115Z, the
             sample must be quantitated by MSA.

     3.  The following  procedures will be incorporated into MSA analyses.

         a)  Data from MSA  calculations must  be within the linear range as
             determined  by  the  calibration  curve generated at the beginning of
             the  analytical  run.
•Spikes are post dlgesc spikes (to be prepared prior to analysis by adding a
 known quantity of the analyte to an aliquot of the digested sample.  The
 unspiked sample aliquot must be compensated for any volume change in che spike
 samples by addition of deionized water to the unspiked sample aliquot.

"''"Spike" is defined as (absorbance or concentration of spike sample) minus
 (absorbance or concentration of the saaple).

-------
          b)  The sample and three spikes oust be analyzed consecutively for
              MSA quantisation (the "initial" spike run data is specifically
              excluded frotn use la the MSA quantication).  Only single injec-
              tions are required for MSA quantitation.

          c)  Spikes* should be prepared such that:

              •  Spike 1 ia approximately 502 of the sample absorbance.

              -  Spike 2 is approximately 100Z of the sample absorbance.

              -  Spike 3 is approximately 1502 of the sample absorbance.

          d)  The data for each MSA analysis should be clearly identified in
              the raw data documentation along with the slope, intercept and
              correlation coefficient (r) for the least squares fit of the data
              and Che results reported on Form VIII.  Reported values obtained
              by MSA are flagged on the data sheet (Fora I) with the letter "V".

          e)  If the correlation coefficient (r) for a particular analysis is
              less than 0.995 the MSA analyses oust be repeated once.  If the
              correlation coefficient is still <0.995, the results on Form I
              oust be flagged with "+".


9.   .Laboratory Control Sample Analysis

     Laboratory Control Sample (LCS) - Aqueous and solid laboratory quality
     control samples must be analyzed for each analyte using the saae sample
     preparation and analytical methods employed for the EPA samples received.

     The aqueous LCS solution must be obtained from EPA (if unavailable, the
     EPA Initial Calibration Verification solution may be used).'  The aqueous
     LCS must be prepared and analyzed with the samples for each of the pro-
     cedures applied to each case of samples received.  One aqueous LCS muse
     be analyzed for every 20 samples received, or for each batch* of samples
     digested whichever is more frequent (see Exhibit D).  Each data package
     must contain the results of all the LCS analyses associated with the
     samples on that Case.  For cyanide, the distilled mid-range calibration
     standard may be used as the aqueous LCS (see Section 8.3.2.1, Exhibit D).
     An aqueous LCS is not required for mercury analysis.
*Spikes are post digest spikes to be prepared prior to analysis by adding
 a known quantity of the analyse to an aliquot of the digested sample.  The
 unspiked sample aliquot must be compensated for any volume change in the
 spiked samples by addition of deionlzed water to the unspiked sample aliquot,

+A group of samples prepared at the same time.
                                      .8-83

-------
All aqueous LCS results will be reported on Form VII in terms of  true
concentration and percent recovery (ZR) as calculated by:

                      ZR • (Observed/True) x 100

where "observed" is the measured concentration.  If the Z  recovery  far
the aqueous LCS falls outside the control limits of 80Z -  120%, the
analyses must be terminated, the problems corrected and the previous
samples associated with that LCS re-analyzed (ie., previous 19 samples
or the batch of samples from the case).

Once a month, a solid LCS, available from EMSL-LV oust be  prepared  and
analyzed using each of the procedures applied to the solid samples  received.
If this EPA solid LCS is unavailable, other EPA Quality Assurance Check
samples or other certified materials may be used.

The monthly results of the solid LCS samples should be reported on  a
duplicate Form VII and submitted monthly to EMSL/Las Vegas and SMO  on
the 15th of every month.

If the percent recovery for the solid LCS sample is outside the control
limits established by EPA, no further sample analyses may  be done until
the analytical problems are solved, and satisfactory LCS results  are
obtained.
                                  5-84

-------
          EXHIBIT G




CHAIN-OF-CUSTODY AND DOCUMENT




      CONTROL PROCEDURES
                B-85

-------
                                     NOTES
The contractor shall not deviate from the procedures described herein without
Che prior written approval of the Contracting Officer:  Provided, that the
Contracting Officer nay ratify in writing such deviation and such ratification
shall constitute the approval required herein.
                                        B-86

-------
                    Specificaclons for Chain-of-Cuscody and
                          Document Control Procedures


The contractor oust have written standard operating procedures (SOP) for
receipt of samples, maintenance of custody, tracking the analysis of samples
and assembly of completed data and must follow these SOPs.  These procedures
are necessary to ensure that analytical data collected under this contract
are acceptable for use in EPA enforcement case preparations and litigation.
The Contractor's SOPs shall provide mechanisms and documentation to meet each
of the following specifications and shall be used by EPA for the basis for
laboratory evidence audits.

1.   The contractor shall have a designated sample custodian responsible for
     receipt of the samples.

2.   The contractor shall have written SOPs for receiving and logging in of
     the samples.  The procedures shall include documentation of the sample
     condition, maintenance of custody and sample security and documentation
     of verification of sample tag information against custody records.

3.   The contractor shall have written SOPs for maintenance of the security
     of the sample after  log  in and shall demonstrate security of the sample
     storage and laboratory areas.

4.   The contractor shall have written SOPs for tracking  the work performed on
     any particular sample.   The  tracking system shall include standard data
     logging formats, logbook entry procedures and a means of controlling log-
     book  pages, computer printouts,  chart  tracing and other written or printed
     documents  relevant to the samples.  Logbooks, printed forms or other writ-
     ten documentation must be available  to describe the  work performed in  each
     of the following stages  of analysis:
          o

     o Sample receipt
     o Sample  preparation
     o Standard preparation/tracking
     o  ICP analysis
     o Flame,  flameless  and  cold vapor  AA  analysis
     o Non-metal  Inorganics  analysis
     o Data  reduction
     o Uata  reporting

 5.   The  contractor  shall have  written SOPs for  organization  and assembly
     of all  documents  relating  to each tPA case.   Documents  shall  be  filed
     on a  case specific  basis.   The  procedures must  ensure that  all documents
      including logbook  pages, sample  tracking records,  measurement  readout
      records,, computer  printouts, raw data summaries,  correspondence  and  any
      other written documents  having  reference to the case, are compiled in
     one  location  for submission to  EPA in the Complete Case  File  Purge package
      (formerly termed the "Document  Control and  Chain-of-Custody Package.")
      The  Case File Purge package must be compiled within 7 days of  submission
    .  of  che sample data package.The system must include a document  numbering
     and  Inventory procedure.
                                      B-87

-------
6.   The Complete Case File Purge package includes but is not limited to:
     sample tags, custody records, sample cracking records, analyses log-
     book pages, bench sheets, measurement readout records, extraction and
     analysis chronicles, computer printouts, raw data summaries, instrument
     logbook pages, correspondence, and the document inventory*
                                   B-88,

-------
                         Hazardous Waste Disposal Sice
                          Contract Analytical Support
                Chain-of-Custody and Document Control Procedures


                                 Sample Control


A sample is physical evidence collected from a facility or from the environment.
An essential part of this investigations effort is the control of the evidence
gathered.  To accomplish this, the following chain-of-custody and document  •
control procedures have been established.

Sample Identification

Each sample bottle shall be labeled with a tag containing the sample number and
sample, description to identify the contents of the bottle.  Additionally, the
sample number shall be marked on the outside of any special packaging container
to facilitate identification.  Typical sample tags are shown in Figure 1.

Chain—of-Custody Procedures

Because of the nature of the data being collected., the possession of samples
must be traceable from the time the samples are collected until they are intro-
duced as evidence in legal proceedings.  To maintain and document sample custody,
the chain-of-custody procedures described here are followed.

A sample is under custody if:

1.   It is in your actual possession, or

2.   It is in your view, after being in your physical possession, or

3.   It was in your possession and then you locked or sealed it up to prevent
     tampering, or

4.   It is in a secure area.

To assure custody of samples during transport and shipping, each sample within
a packing container is recorded on a chain-of-custody record shown in Figure 2.
Each sample number Is recorded and the number of containers shipped is recorded
on the sheets.  Also record the other information regarding the project, sam-
ples (or shipper if returning empty bottles), method of shipment and remember
to sign and date the sheet.  The original custody sheet is then placed inside
the package (protected from damage) and the package sealed.  Sample containers,
shipping boxes, coolers or other packages may be sealed by using the seal shown
in Figure 3.  The seal must be placed so the container cannot be opened without
breaking the'seal.
                                         B-89

-------
 Upon receipt of samples  in custody,  Inspect the package and note any damage co
 the sealing tape or custody seals.  Note on the custody record or other log-
 book that  the seals or locks were intact upon receipt if no tampering or damage
 appears to have occurred.   Open the package and verify that each item listed on
 the sheet  is present and correctly identified.  If all data and samples are cor-
 rect, sign and date the  "received by Laboratory by" box.  In the event errors
 are noted,  record the discrepancies  in the remarks column (initial and date each
 comment) then sign the chain-of-custody record.  Report discrepancies to the
'Sample Management Office for remedies.

 Laboratory Document Control

 The goal of the Document Control Program is to assure that all documents for
 a specified case (group  of samples)  will be accounted for when the project is
 completed.   The program  Includes a document numbering and inventory procedure
 for preparation of the specified documentation packages for each case.

 Logbooks

 All observations and results recorded by the laboratory but not on preprinted
 data sheets are entered  Into permanent laboratory logbooks.  Data recorded are
 referenced  with the case number, date and analyst's signature at the top of the
 page.  Data from only one case are recorded per page.  When all the data from a
 case is compiled, copies of all logbook entries must be included in the documen-
 tation package.

 Instrument  logs are also limited to  one case per page with the ease number
 recorded at the top of each page.  Copies of these logs must also be included
 in the final documentation package.

 Corrections to Docuaentation

 All documentation in logbooks and other documents shall be in ink.   If an
 error Is made in a logbook assigned  to one individual, that person should make
 corrections simply by crossing a line through the error and entering the correct
 information.  Changes made subsequently are dated and initialed.  Corrections
 made to other daca records or non-personal logbooks are made by crossing a
 single line through the  error, entering the correct information and initialing
 and* dating  the correction.

 Consistency of -Documentation

 Before releasing analytical results,  the laboratory assembles and cross checks
 the information on sample tags, custody records, lab bench sheets,  personal and
 instrument  logs andother relevant data to ensure that data pertaining to each
 particular sample or case Is consistent throughout the record.

-------
Doeuaene Numbering  and  Inventory Procedure

In order Co  provide docuoenc  accouncabilicy  of  che  completed  analysis  records,
each Item in a  case is  inventoried and assigned a serialized  number and an
identifier associating  it TO  the case and Region.

           Case # - Region -  Serialized number

           For  example  - 75-2-0240

All documents relevant  to each case including:  logbook pages, bench sheets,
instrument readings, chare recordings, custody records, etc., are inventoried.
Each data generator (analyst) is responsible for ensuring that all documents
generated are placed in the file for inventory and  returned to EPA.  Figure 4
is an example of a  document inventory*

Confidential Information

Any samples  or  information received with a request  of confidentiality is handled
as "confidential."  A separate, locked file  is maintained and segregated from
other nonconfidential information.  Data generated  from confidential samples is
treated as "confidential."  Upon receipt of  confidential information, the DCO*
logs these documents into a Confidential Inventory  Log.  The information is
then made available to  authorized personnel, but only after it has been signed
out to that person  by the DCO.  The documents shall be returned to the locked
file at the  conclusion  of each working day.  Confidential information may not
be reproduced except upon approval by and under the supervision of the DCO.
Any reproduction must be kept to an absolute minimum.  The DCO will enter all
copies into  the document control system and apply the same requirements as
the original.   In addition, this Information may not be destroyed except upon
approval by and under the supervision of the EPA program manager and EPA con-
tracting officer.   The  DCO shall remove and  retain  the cover page of any con-
fidential information disposed of for one year and  shall keep a record of che
destruction  in  the  Confidential Inventory Log.
*DCO is a document concrol officer assigned by che contractor to maintain
 control of confidential information.

-------
     tf-l'f'J
       yoo?
*

MI:
*!•* I Hi'
                     I
.§
                         !
                         i
           v>EPA
   UNfTED STATES ENVIRONME^AL PROTECTION AGENCY
     Figure 1.  Example sample tag.
                  B-92

-------
             r*K'mONMINtA|

                    O" ** !•' I nliM
                                    AGf NCV
                                                           CHAIN OF CUSTOOV Rf COHO
   oc
   c
00 3
 I  I
vo o
to *»>

   r.
   c
   o:
   n
   o
rncij NO f
ntuiri MAMr


SAI''PII"S 	 '
««• no
- -













„.„






-•







IIMI
•_
-•


	





q...n,u.,l.,db, ,».,-.,...



1
II
—
--





...

4

-•


-
--

—


	 -








	 	









D«lr / Tim*
D.i./rm,
"~P
"•»--•••» 	 -V

NO
CON
IAIN! MS


	











Rtc*iv*d by ISfpMiM**! t
Rrr* iv»d by 1 S*p*iwr«l
R»c*i**«l >n* ItltQtftinty hf
/////// ««»»M

•































"





































































•

Hd.nqu.iKtdbv IS^M.^ O.It f.B- RK.IMd b* n.*«~~>


Dal
, »«..»..! D«t T.IM Ht».rtdbV 0-r..*^.
l/limt *
., — , r..:..r.» 	 	
UiMiki

-------
"IV3S AQOJLSnO
  CUSTODY SEAL
L     -	-
  Date -
  Signature
  Figure 3.  Example custody seal made of perforated paper stock.

-------
1.   Scudy plans or project plans.

2.   Saaple traffic records, weekly reports.

3.   Custody records, sample tags, sample loop.

4.   Laboratory logbooks, personal logbooks, instrument logbooks (or
     appropriate copies of logbook pages).

5.   Laboratory data (sorted by sample), calibration and quality
     control data results.

6.   Data summaries and reports.

7.   All other documents, forms or records referencing the samples.
          Figure 4.  Example Document Inventory Format for each case*.
                                        B-95

-------
            APPENDIX C



ELUTRIATE AND FRACTIONATION METHODS



            (PLUMB  1981)

-------
Elutriate test
           The elutriate test is a simplified simulation of the
dredging and disposal process wherein predetermined amounts of
dredging site water and sediment are mixed together to approximate
a dredged material slurry.2'20  The elutriate in the supernatant
resulting from the vigorous 30-min shaking of one part sediment from
the dredging site with four parts water (vol/vol) collected from
the dredging site followed by a 1-hr settling time and appropriate
centrifugation and O.U5 v filtration.  Thus, it will be necessary to
collect both water and sediment samples to perform the elutriate test.
When evaluating a dredging operation, the sediment should be collected
at the dredging site and the water should be collected at the dredging
and the disposal site.  To evaluate a fill material activity, samples
should be collected from the source of the fill material and the water
should be collected from the disposal site.
           Water sample collection.  Collection should be made with an
appropriate noncontaminating water sampling device.  Either discrete
samplers such as Kemmerer or Van Dorn samplers or continuous collectors
such as submersible pumps may be used.  The volume of water required
will depend on the number of analyses to be performed.  For each sample
to be subjected to elutriate testing, it is suggested that a minimum of
It £ be collected at the disposal site and 8 i be collected at the
dredging site to evaluate a dredging operation and/or 12 i be collected
at the disposal site to evaluate a fill material disposal operation.
This will provide li i of water for analyses and sufficient water to
prepare triplicate 3-& elutriates.  (Each elutriate should yield
2.0 to 2.2 i of standard elutriate for analysis.)  If the samples are
to be analyzed for trace organics or a large number of constituents,
a proportionately larger initial sample should be collected.
           Samples must be stored in glass containers if trace
organic analyses are to be performed.  Generally, either plastic or
                                         C-l

-------
glass containers may be used for other parameters.  The samples should
be maintained at U°C until analyzed but never frozen.  The storage
period should be as short as possible to minimize changes in the
characteristics of the water.  Disposal site water should be analyzed
or split and preserved immediately.  The remainder of the water should
be used in the elutriate test, which should be processed within 1 week
of collection.
           Sediment sample collection.  Samples should be taken from
the fill or the dredging site with a grab or a corer.  Approximately 3&
of sediment or fill material would provide sufficient sample to prepare
triplicate 3-i elutriates.  Again, if the resultant standard elutriates
are to be analyzed for trace organics or a large number of constituents,
a proportionately larger initial sample should be collected.
           Samples may be stored in plastic bags, jars, or glass
containers.  However, if trace organic analyses are to be performed,
glass containers with teflon-lined lids are required.  A special
precaution that must be taken with sediment samples is to ensure
that the containers are completely filled with sample and that air
bubbles are not trapped in the container.  This step is necessary to
minimize sample oxidation that could influence elutriate test
results.2'22
           The samples should be stored immediately at U°C.  They must
not be frozen or dried prior to use.  The storage period should be as
short as possible to minimize changes in the characteristics of the
sediment.  It is recommended that samples be processed within 1 week
of collection.
           Apparatus.  The following apparatus are required to perform
the elutriate test.  Prior to use, all glassware, filtration equipment,
and filters should be washed with 5 to 10 percent (or stronger)
hydrochloric acid (HCl) and then rinsed thoroughly with deionized
water.  The necessary apparatus include:
           a.  Acid-rinsed plastic bottles for collection of water
               samples.
           b.  Plastic Jars or bags ("Whirl-Pak," plastic freezer
               container's, etc.) for collecting dredged or fill
               material samples.
                               C-2

-------
c_.  Laboratory shaker capable of shaking 2-i flasks at
    approximately 100 excursions/minute.  Box type or
    wrist-action shakers are acceptable.

d_.  Several 1-i graduated cylinders.

e_.  Large (15 cm) povder funnels.
£.  Several 2-1, large-mouth graduated Erlenmeyer flasks.

£.  Vacuum or pressure filtration equipment, including
    vacuum pump or compressed air source, and an appropriate
    filter holder capable of accomodating U7-, 105-, or
    155-m-diameter filters.

h.  Membrane filters with a O.U5-V pore-size diameter.
~"   The filters should be soaked in 5 M HC1 for at least
    2 hr prior to use.
i_.  Centrifuge capable of handling six 1- or 0.5-A centri-
    fuge bottles at 3000 to 5000 rpm.  International Model
    K or Sorval Super Speed are acceptable models.

J_.  Wide-mouth, 1-gal capacity glass Jars with teflon-
    lined screw-top lids for use as sample containers
    when samples are to be analyzed for trace organics.
    (It may be necessary to purchase Jars and teflon
    sheets separately; in this case, the teflon lid
    liners may be prepared by the laboratory personnel.)
Test procedure.  The stepwise test procedure is given below:

a.  Subsample a minimum volume of 1 I each of dredging site
    and disposal site water.  If it is known in advance
    that a large number of measurements are to be performed,
    the size of each subsample should be increased to meet
    the anticipated needs.

b.  Filter an appropriate portion of the disposal site
~   water through an acid-soaked 0.1*5-y pore-size membrane
    filter that has been prerinsed with approximately
    100 ml of disposal site water.  The filtrate from the
    rinsing procedure should be discarded.
£.  Analyze the filtered disposal site sample as soon as
    possible.  If necessary, the samples may be stored at
    U°C after splitting and the appropriate preservatives
    have been added (Table 2-U).  Filtered water samples
    may also be frozen with no apparent destruction of
    sample integrity.
d_.  Repeat steps a., b, and c_ with dredging  site water.
    This step is omitted with a fill material sample.
e.  Subsample approximately 1 iof sediment  from the well-
~~   mixed original sample.  Mix the sediments and
    unfiltered dredging site water  in a volumetric
    sediment-to-water ratio of 1:U  at room  temperature
    (22 +_ 2°C).  This is best done  by the method of
                          C-3

-------
                volumetric displacement.23  One hundred mililiters  of
                unfiltered dredging site water is placed into a
                graduated Erlenmeyer flask.   The sediment subsample is
                then carefully added via a powder funnel to obtain  a
                total volume of 300 ml.   (A  200-ml volume of sediment
                will now be in the flask.)  The flask is then filled
                to the 1000-ml mark with unfiltered dredging site
                water, which produces a slurry with a final ratio of
                one volume sediment to four  volumes water.

            This method should provide 700 to 800 ml of water for

analysis.  If the analyses to be run require a larger volume of water,
the initial volumes used to prepare the elutriate slurry may be
proportionately increased as long as the solid-to-liquid ratio remains

constant (e.g. mix UOO ml sediment and 1600  ml unfiltered dredging

site water).  Alternately, several 1-i sediment/dredging site water
slurries may be prepared as outlined above and the filtrates combined

to provide sufficient water for analysis.  The procedure continues  as

follows:
            £.  (1)  Cap the flask tightly with a noncontaminating
                     stopper and shake vigorously on an automatic
                     shaker at about 100 excursions per minute for
                     30 min.  A polyfilffl-covered rubber stopper is
                     acceptable for minimum contamination.

                (2)  During the mixing step given above, the
                     oxygen demand of the dredged material may cause
                     the dissolved oxygen concentration in the elutriate
                     to be reduced to zero.   This change can alter
                     the release of chemical contaminants from dredged
                     material to the disposal site water and reduce
                     the reproducibility of the elutriate test. l
                     If it is known that anoxic conditions (zero
                     dissolved oxygen) will not occur at the disposal
                     site or if reproducibility of the elutriate test
                     is a potential problem, the mixing may be
                     accomplished by using a compressed air-mixing*
                     procedure instead of the mechanical mixing
                     described in Step f. (l).  After preparation of
                     the elutriate  slurry, an air-diffuser tube is
                      inserted almost to  the bottom of the flask.
                      Compressed air should be passed through a
                      deionized water trap and then through the
                      diffuser tube  and the slurry.  The flow rate
                      should be adjusted  to agitate the mixture
 *  This procedure can  cause the loss of highlj volatile chemical con-
   stituents.   If volatile materials are of concern, compressed air
   mixing  should not be used.
                               C-4

-------
                     vigorously for 30 min.   In addition,  the  flasks
                     should be stirred manually at  10-min  intervals
                     to ensure complete mixing.

            £.        After 30 min of shaking or mixing with air,
                     allow the suspension to settle for 1  hr.

            li.        After settling, carefully decant  the  supernatant
                     into appropriate centrifuge bottles and then
                     centrifuge.  The time and revolutions per minute
                     during centrifugation should be selected  to  reduce
                     the suspended solids concentration substantially
                     and, therefore, shorten the final filtration
                     process.  After centrifugation, vacuum or pressure
                     filter approximately 100 ml of sample through a
                     O.U5-U membrane filter and discard the filtrate.
                     Filter the remainder of the sample to give a
                     clear final solution (the standard elutriate) and
                     store at U°C in a clean, noncontaminating container
                     in the dark.  The filtration process  is intended
                     for use when the standard elutriate is to be analyzed
                     for conventional chemical contaminants.  When the
                     elutriate is to be analyzed for organic contaminants
                     and PCB's, filtration should not be used  since
                     organic concentrations can be  reduced by  sorption.
                     Centrifugation should be used  to remove particulate
                     matter when the standard elutriate is to  be
                     analyzed for specific organics.
            i_.        Analyze the standard elutriate as soon as possible.
                     If necessary, the samples may  be stored at U°C
                     after splitting and the appropriate preservatives
                     have been added.
            j_.        Prepare and analyze the elutriate in triplicate.
                     The average of the three replicates should be
                     reported as the concentration of the standard
                     elutriate.

Sediment fractionation
            Chemical constituents associated with sediments may be

distributed in many chemical forms.  The purpose of a fractionation
procedure is to better define this distribution.  This objective is
achieved by leaching a sample with a series of successively harsher

extraction agents.  Reagents used in the procedure to be described

below consist of interstitial water, ammonium acetate, hydroxylamine,

hydrogen peroxide, citrate-dithionate, and hydrofluoric acid-nitric

acid.
            The premise of the  fractionation procedure  is that a

specific geochemical phase is defined by a  specific chemical


                                         C-5

-------
extraction agent.  Thus, the ammonium acetate extract is referred to as
the exchangeable phase, and the citrate-dithionate extract is referred
to as the moderately reducible phase.  These relationships have not been
rigorously demonstrated and, therefore, the fractions are only opera-
tionally defined.
            The use of fractionation results at this time appears to be
limited to supporting other studies.  That is, results have greater
value in sediment research studies than in the regulatory decision-
making process.  Limited results with sediment fractionation data
suggest a higher correlation vith the more labile sediment phases
(interstitial water, exchangeable phase) and elutriate test results11
and long-term water quality changes.l°
            A major limitation of the fractionation procedure is that
previous experience is limited to heavy metals and nutrients.  A broad
spectrum analysis of the individual fractions has been limited by small
sample size, particularly the interstitial water fraction.
            Sample collection.  Samples should be collected with a grab
or a corer.  Because the distribution of sediment-associated chemicals
can be altered by processes such as drying and oxidation, samples should
be kept wet and exposure to the atmosphere should be minimized.  Samples
collected with a grab or dredge must be quickly transferred to a con-
tainer and air bubbles must be excluded from the container.  Corer
samples should be sealed in the core liner and returned to the labora-
tory in an upright position.  Previous studies have shown that a 15-cm
section from a 7.5-cm-diamcter core can provide sufficient material to
perform fractionation studies for eight metals and four nutrients.11
            The samples should be stored immediately at U°C.  They
must not be frozen or dried prior to use.  The storage period should
be as short as possible to minimize changes in the distribution of
chemical constituents in the sediments.  It is recommended that samples
be processed within 1 week of collection.
            Apparatus.  The following apparatus is required to perform
the elemental partitioning procedure.  Prior to use, all glassware,
filtration equipment, and filters should be washed with 5 to 10 percent
HC1 and then rinsed thoroughly with deionized water.  This list of
                               C-6

-------
equipment includes:
            a.  Plastic Jars or bags ("Whirl-Pak,"  plastic  freezer
            ~   containers,  etc.)  for collecting grab or  dredge
                samples or polyethylene liners for  the collection of
                core samples.
            b_.  Glove box or disposable glove bag.
            £.  Polarographic oxygen analyzer or alternate  method to
                confirm the absence of oxygen in glove bag.
            £.  Utensils for splitting cores and handling samples in the
                glove bag.
            e_.  250-ml and 500-ml polycarbonate centrifuge  bottles.
            f_.  Refrigerated centrifuge.
            £.  Vacuum filtration apparatus.
            h.  150-ml and 120-ml polyethylene storage bottles.
            i_.  Blender or porcelain mortar and pestle.
            j_.  Top-loading balance.
            k.  Weighing dishes.
            1^  Digestion block or hot plate.
            m.  Teflon beakers.
            la.  50-ml volumetric flasks.
            Test procedure.  A stepwise sediment fractionation procedure
is given in the following paragraphs.
            To begin the procedure, prepare a glove box or disposable
glove bag.  Flush the system with nitrogen gas and maintain a positive
pressure nitrogen atmosphere.  Oxygen-free conditions in the glove bag
or box should be verified with a polarographic oxygen analyzer prior
to sample processing.  Initial  sample handling and all steps in the
interstitial water and ammonium acetate extractions  should be conducted
under a nitrogen atmosphere.
            Acid wash all hardware  to be used in the extractions  in
6 N HCL and thoroughly rinse with distilled water to minimize  sample
contamination during processing.
            The initial  separation  removes the  interstitial water phase.
This is accomplished by  first  placing  the sealed sediment  sample  in the
glove bag.  After reestablishing the nitrogen atmosphere,  extrude the
sediment core from  its liner into a flat plastic container.  If  the
core is to be sectioned  vertically, 15-cm sections of a  7.5-cm-diaraeter

                                         C-7

-------
core have been shown to provide sufficient material for the sequential
fractionation procedure.  Each core section should be split into halves
vith one half (approximately 300 cc) being used for the interstitial
vater testing and the remaining half used for all other analyses.  Place
the half section for the interstitial water analysis in an oxygen-free,
polycarbonate 500-ml centrifuge bottle in the glove bag and seal.
Centrifuge the sample in a refrigerated centrifuge (U°C) at 900
revolutions per minute (M.3,000 * g) for 5 min.  This should be
sufficient to recover UO percent of the total sediment water.  After
centrifugalion, return the sample to the glove bag and vacuum filter
the interstitial water through a O.U5-V pore-size membrane filter.
Transfer the filtered sample to an acid-washed polyethylene bottle and
acidify to pH 1 with HC1 for preservation.
            If the sediment sample is not to be sectioned vertically,
decant excess water and blend the core or dredge sample.  Place
approximately 300 cc of the blended sample in an oxygen-free, poly-
carbonate 500-ml centrifuge bottle and seal.  Centrifuge this sample
for 5 min at 900 revolutions per minute (I3j000 * g) in a refrigerated
centrifuge (U°C) and then filter through a O.U5-U pore-size membrane
filter under a nitrogen atmosphere.  The filtered sample may be
analyzed immediately or split for preservation and storage.*
            The exchangeable phase is determined on the unused half of
the wet sediment sample that was blended for interstitial water
analysis.  Blend the wet sediment with an electrically driven poly-
ethylene stirrer contained in the glove bag.  Remove a subsample of
the homogenized sediment sample (blended core section, core, or grab
sample) for percent solids determination.
            Weigh a second subsample (approximately 20 g dry weight
of each homogenized sediment section into an oxygen-free, tarred,
250-ml centrifuge tube containing 100 ml deoxygenated 1 N^ ammonium
acetate, producing a suspension with an approximate solid-to-liquid
ratio of 1:5.  Adjust the pH of the acetate solution to the pH of the
surface sediments.  Seal the samples and then place on a wrist-action
shaker for 1 hr.  Centrifuge the samples at 6000 revolutions per minute
for 5 min and return them to the glove bag for further processing under
                              C-8

-------
the nitrogen atmosphere.  Filter the sample through O.U5-V  pore-size
membrane filters, retaining both the filtrate and the solid residue.
            The filtrate may be analyzed immediately or split  and
preserved for specific constituents, as discussed for water samples.
This extract will also include the interstitial water components
since a fresh blended sediment sample was used.  Therefore* measured
concentrations should be reduced to compensate for the interstitial
water.  This can be accomplished as:

            Wt exchangeable material = (Vol ext) (Cone ext) -
                    (Wt Sample) (l-< Solids) ,   .
                        (density water)      uw<;
where
(Vol ext)       = volume of ammonium acetate extract
(Cone ext)      = analytical concentration in ammonium acetate extract
(Vt Sample)     = wet weight of sample for ammonium acetate extraction
% Solids        = percent solids in sample
(density water) = density of water at temperature of sample
(IWC)           = interstitial water concentration of sample
            The easily reducible phase is performed with the solid
residue from the exchangeable phase determination.  This step and all
subsequent steps in the fractionation procedure can be conducted outside
the glove bag.  Add 50 ml 1?2 sparged distilled-deionized water to the
centrifuge tube containing the solid residue from the 1 N. ammonium
acetate extraction.  Agitate the sample with a stainless steel spatula
or a glass-stirring rod to ensure good washing efficiency.   Centrifuge
the suspension of 6000 revolutions per minute and discard the  liquid
phase.  A portion of the solid residue will have to be set  aside at
this point for a redetermination of percent solids.
            Blend the remaining sediment residue and transfer a 2-g
(dry weight equivalent) subsample to a 250-ml Erlenmeyer flask.  Add
100 ml of 0.1 M hydroxylamine hydrochloride-0.01 M nitric acid solution.
The resultant suspension will have a solid-to-extractant ratio of
approximately 1:50.  Seal the sample and place the suspension on a
wrist-action shaker (or equivalent) for 30 min.  Centrifuge the sample

                                         C-9

-------
at 6000 revolutions per minute for 5 min.  Decant and filter the liquid
phase through 0.1»5-U pore-size membrane filters.  The filtrate may be
treated as a water sample and analyzed immediately or split and preserved
for specific constituents.
            The solid residue is used in the organic and sulfide phase
extraction.  Wash the residue from the easily reducible phase vith 50 ml
distilled water.  After agitating the suspension, centrifuge the sample
at 6000 revolutions per minute for 5 min and discard the supernate.
                                *
Subsample the residue for a percent solids determination so the organic
and sulfide results can be expressed on a dry weight basis.  Add 50 ml
of 30 percent hydrogen peroxide to the washed residue and adjust the pH
to 2.5 with HC1.  (The purpose of the pH adjustment is to prevent any
released metals from precipitating.)  Digest the sample at 95°C for
6 to 8 hr.  Add 100 ml of 1 If ammonium acetate buffered at pH 2.5 to
the digestate and shake for 1 hr.   Centrifuge the sample at 6000 revo-
lutions per minute for 5 min and filter the sample through 0.1* 5-w
pore-size membrane filters.  Treat the filtrate as a water sample
and analyze immediately or split and preserve as required.  Retain
the solid residue.
            The next fraction in the 'sequence is the moderately
reducible phase.  Wash the organic and sulfide phase solid residue
with 50 ml of distilled water; centrifuge as described earlier; and
discard the supernate.  Redetermine percent solids on a subsample
of the residue.  Add 100 ml of a citrate-dithionate solution
(16 g sodium citrate + 1.67 g sodium dithionate/100 ml distilled
water) and mechanically shake the suspension for 17 hr.  Centrifuge
the sample at 6000 revolutions per minute for 5 min and filter the
supernate through a O.b5-V pore-size membrane filter.  Analyze the
filtered sample for moderately reducible constituents and retain
the solid residue for further treatment.
            The sample for residual phase digestion is obtained by
washing the moderately reducible phase residue with 50 ml of distilled
water; centrifuging at 6000 revolutions per minute for 5 min; and
discarding the supernate.  Dry the residue at 105°C and transfer
                              C-10

-------
a 0.5-g dry weight subsample to a teflon  beaker.  Add 15 ml hydro-
fluoric acid and 10 ml concentrated nitric acid;  cover  the beaker;
and digest at 175°C.  After evaporation to near dryness, add  8 ml
fuming nitric acid stepwise in 2-ml increments.  Continue evaporation
to near dryness.  Add 6 N. HC1 to dissolve the residue,  heating if
necessary.  Quantitatively transfer the solution  to a 50-ml volumetric
flask and dilute to volume.  Analyze the sample immediately or preserve
subsamples for specific constituents.
            A schematic flov diagram for the fractionation procedure
is presented in Figure 2-2.  When this procedure  is used, a built-in
quality control check is to total each of the operationally defined
phases and compare to a total digest of the sample. The data should
be considered suspect if they differ by more than 5 to  10 percent.
Bulk analysis
            A bulk analysis provides a measure of the total concen-
tration of a specific constituent in the sample being analyzed.  This
is accomplished by subjecting a sample to strong  oxidation, acid
digestion, or organic solvent extraction.  The procedure is similar
to that used for the residual phase digestion in  the elemental
partitioning procedure discussed earlier.  Total  sediment concentrations
can be used to compare different sites and to identify  major  point
sources.  However, because of the harshness of the extraction pro-
cedure, information on chemical distribution and/or potential
environmental impact is lost.
            Sample collection.  Samples for total analyses may be
collected with a dredge or grab sampler or a core sampler.  Approxi-
mately 1 to 2 I of sediment or fill material should be taken  from the
proposed project site and placed in plastic Jars  or containers.   If
trace organic constituents are to be determined,  the sample  should be
stored in a glass container or a second sample of approximately the
same size should be collected and stored separately in glass  containers.
Samples should be stored at U°C.
            Upon reaching the laboratory, sediment samples may be
stored wet, air dried, or frozen.  The selection between these
preservation techniques should be'based primarily on the specific
                                        C-ll

-------
I
t-»
ro
NTERSTITIAL WATER
     PHASE
SEDIMENT SAMPLE I
   (Preserve
    Anaerobic
    Integrity)
   Section and
   Centrifuge
        •^RESIDUE
  upernatant
    (Filter)
    Acidify
    (pH  1-2)

 SOLUTION

 ANALYZE AS
   INTERSTITIAL
   WATER PHASE
                              EXCHANGEABLE PHASE
                              EXTRACT with
                                  IN NHtOAc
                                  -Centrifuge
                                  (Preserve
                                  Anaerobic
                                   Integrity)
                          Wast
                               Supernatant
                                  (Filter)
                                  Acidify
                               SOLUTION

                               ANALYZE AS
                                 EXCHANGEABLE
                                 PHASE
EASILY REDUCIBLE
     PHASE
EXTRACT wi th
   NH2OH-HC1
                                                      Centrifuge
Supernatant
"   (Filter)
SOLUTION

ANALYZE AS
   EASILY
   REDUCIBLE
   PHASE
  ORGANIC +
SULFIDE PHASE
DIGEST with
       H202
   (pH 2.5)
   at 95°C
                                                           Extract as
                                                               Exchange-]
                                                               able
                                                               (pH 2.5)
                                                                 •RESIDUE
Supernatant
    (Filter)
SOLUTION

ANALYZE AS
  ORGANIC +
  SULFIDE
  PHASE
  MODERATELY REDUCIBLE
           PHASE
EXTRACT with Na2S20H
                                                                                                      RESIDUAL PHASE
DIGEST at 95°C
      fh HF +
      , +
      ming)
      s
                                                                                       entrifuge
                                                                                                 Wash
                                                     IESIDUE
Supernatant
I   (Filter)
SOLUTION

ANALYZE AS
  MODERATELY
  REDUCIBLE
  PHASE
                                              (Centrifuge)
I
SOLUTION

ANALYZE AS
  RESIDUAL
  PHASE
               Figure 2-2.     Elemental  partitioning for sediment characterization

-------
parameter to be determined and,  secondarily,  on personal  choice.

Several parameters such as pH, redox,  total solids,  and volatile

solids must be run on wet samples.   Other parameters may  change due

to oxidation (chlorine demand, BOD,  COD,  SOD, sulfides),  volatili-

zation (phenolics, volatile solids), or chemical instability (carbamates,

herbicides).  Samples to be analyzed for  these parameters should  be

processed as soon as possible using  subsamples of the original vet

sample.  Samples to be analyzed for  particle size (dispersed), total

organic carbon (TOC), metals (except possibly mercury), chlorinated
hydrocarbon pesticides, and PCB's may  be  stored wet, dried,  or frozen.

            Apparatus.  The specific equipment necessary  will vary

depending on the chemical constituent(s)  to be analyzed in the total

sediment digest or the total sediment  organic extract. Specific

needs and cleanup procedures are discussed with each parameter in

Section 3.
            Test procedure.  The following stepwise procedure is

recommended for the processing of sediment or fill material samples

to be analyzed for total or bulk content:
            a_.  Decant any overlying water that may have  been collected
                with the dredge or corer.
            b_.  Blend the dredge, core, or sectioned core sample.

            £.  Transfer an aliquot  of the homogenized sample to a
                tarred weighing dish and weigh.  Dry the  sample at
                105°C to a constant  weight.  This information will
                allow calculation of percent solids in the sample
                and to report subsequent bulk analysis results on a
                milligram-per-kilogram dry weight basis.   The dried
                sample from the percent solids determination may be
                subjected to further chemical analysis for those
                parameters not affected by the drying process.

                If volatile solids are to be determined,  record the
                weight of the crucible and the dried sample used in
                the percent solids determination.  Place the sample
                in an electric muffle furnace and ignite the sample
                at 600°C for 60 min.  Remove the sample from the
                furnace, allow to cool, and desiccate for 30 min
                prior to weighing.  Report the weight lost on
                ignition as percent volatile solids.
            d_.  Transfer a second subsample of the blended sample to
                a suitable container for pH and oxidation-reduction
                (redox) potential determinations.  The sample  size

                                         C-13

-------
                should be sufficient to allow the electrodes to be
                inserted to a depth of U to 6 cm.  Allow sufficient
                time for the electrode responses to stabilize and
                record the respective pH and redox values.

            e_.  Set aside subsamples of the wet, blended sample for the
                analysis of time-dependent or unstable chemical
                constituents.  Parameters in this category include
                biological oxygen demand, chemical oxygen demand,
                sediment oxygen demand, chlorine demand, herbicides and
                carbamates, phenolics, sulfides, nitrogen, phosphorus,
                and oil and grease.  Thus, as many as 12 subsamples
                (if all listed analyses are to be performed) will be
                required.  Suggested sample sizes for each aliquot are
                presented in Figure 2-3.  These analyses should be
                initiated as soon as possible to minimize the effects
                of sample alteration due to handling and storage.

            f_.  Set aside separate subsamples for the analysis of
                particle size, carbon, metals, and chlorinated
                hydrocarbons.  However, because of the increased
                stability of these constituents (relative to those
                in Step £, above), the aliquots may be taken from the
                initial wet, blended sample, or a sample that has been
                dried or frozen for storage.  Required subsample sizes
                are presented in Figure 2-3.

            Individuals performing bulk analysis of sediment samples
should be aware of the fact that analytical results may be affected by

sample handling and storage procedures.  The following special caveats
are maintained here because of the importance of this fact and again

with the appropriate analytical procedure in Section 3:
            a_.  It is preferable to determine Eh and pH values in the
                field as soon as the sample is collected since there is
                no way to stablize these parameters.  If this is not
                possible, these parameters should be determined as
                soon as possible in the laboratory using a wet sample.
                Sample handling should be-kept to a minimum to avoid
                sample dehydration or sample oxidation.
            b_.  Percent solids and specific gravity also must be
                determined on a sample of original moisture content.
                The sample should be handled in such a manner to
                minimize water loss and sample dehydration.
            c_.  Cation exchange capacity can be influenced by sample
                drying.  Therefore, it is recommended that this
                parameter be determined on original moisture content
                samples.
            d_.  Chlorine demand, biological oxygen demand, chemical
                oxygen demand, and sediment oxygen demand are all
                              C-14

-------
            Dredge or Grab
                Sample	
Mineralogical
 Compos i t i on
   10-25 9
    Total
   Organic
    Ca rbon
      &
    Total
   Inorganic
    Carbon
    1-2 q
Heavy Metals
    1-10 g
   PCB's &
  Pesticides
   10-100 q
Time-dependent Parameters
   Process Immediately
Particle Size
   25-50 q
Whole
Core

1 BLEND SAMPLE



^ Dry Sample L


^ freeze Sample L
*"" for storage ™




Note: All sa
a dry


Section
Core


^
^ ^
Dry Sample pH
105 *C Eh
% Solids 50-100 q
5-20 q





600 "C
Volatile Solids
mple sizes given on
weight basis.


— »
_k




— fr
,__*
W
— •+

_fc

— •*
— 1
— 1
-H
—4
Cation Exchange Capacity
5-10 q
Chlorine Demand
1-2 g
Biochemical Oxygen Demand
0.5-5.0 q
Chemical Oxygen Demand
0.5-5.0 q
Sediment Oxygen Demand
10-100 q
Herbicides & Carbamates
10-100 q
(Phenolics
10-50 a
{Sul fides
1-5 a
{ Spec! fie Gravity
250 q
\ Nitrogen forms
t 0.5-10 q
\ Phosphorus forms
T 0.5-10 q
J Oi 1 6 Grease
                                                                                     5-50 g
                      Figure  2-3.    Sediment sample splitting for bulk analysis

-------
    measures of the reducing capacity of the sample being
    analyzed.  Since sediments are frequently reduced and
    contain elevated concentrations of ferrous iron,
    manganous manganese, and sulfide that can be oxidized
    by atmospheric oxygen, these parameters should be run
    on vet samples.
e_.  Herbicides and carbamates are chemically unstable with
    relatively short half-lives.  Immediate extraction of
    the original sample with methylene chloride reduces
    the possibility of chemically or biologically catalyzed
    decomposition and increases herbicide and carbamate
    stability.
£.  Phenolic compounds may be lost by volatilization during
    storage.  Therefore, samples to be analyzed for phenols
    should not be dried and storage time should be minimized.
    If immediate analysis is not possible, storage by
    freezing may be acceptable.  Subsequent sample thawing
    should be accomplished at low temperature to reduce
    phenol loss by volatilization.
£.  Sulfides in the sample may be lost by volatilization
    and oxidation.  It is recommended that sample contact
    with atmospheric oxygen be minimized between sample
    collection and analysis to reduce this effect.  This
    can best be accomplished by excluding air bubbles from
    sample containers and minimizing sample storage time.
h.  Some forms of nitrogen that are expected to occur in
    sediments (nitrites) are unstable in the presence of
    oxygen and can be lost on sample drying.  In addition,
    sample composition may be altered by the uptake or
    loss of volatile ammonia.  Therefore, sample processing
    should begin as soon as possible using a sample of
    original moisture content.
i^  The distribution of phosphorus forms may be altered by
~  changes in other sample constituents.  For example,
    the oxidation of iron in a sample may precipitate
    soluble phosphate.  Therefore, if soluble phosphate
    is to be determined, wet samples should be processed
    as soon as possible.  If total phosphate is the only
    parameter of concern, analysis can be conducted on a
    wet, dried, or frozen sample.
j_.  The oil and grease  content of samples may be  reduced
    due to volatilization.  Consequently, sample  drying
    prior to analysis is not recommended.
k_.  The selection  of a  storage technique for samples to be
    analyzed for particle size depends on the method of
    analysis.  If  apparent particle  size is to be run,
    a wet sample should be used.  However,  if dispersed
    particle  size  is to be run, a wet, dried, or  frozen
    sample may be  used.
                   C-16

-------
            1^  Most of the heavy metals are stable and samples
                scheduled for analysis can be stored in a vet, dried,
                or frozen state.  The selection of a storage method
                can affect the distribution of a metal among various
                forms, but the total concentration should be unaffected.
                Tvo possible exceptions are mercury and selenium, which
                can be lost by volatilization.  This particularly is
                true if the samples are dried above 60°C.

            m.  Chlorinated hydrocarbon pesticides and PCB's are stable
                and probably unaffected by the method of sample  storage.
                Improved stability can be achieved by immediate
                extraction of the original sample with an organic
                solvent and is suggested, but not essential.
            11.  The analysis of a sediment or fill material sample  for
                specific constituents will require a sample digestion
                or sample extraction technique.  Since the selection of
                a digestion solution or solvent is dependent on  the
                analysis to be performed, this information is presented
                with the specific analytical techniques.
            Dredged material may be subjected to several types of

testing.  This section has provided guidance for conducting elutriate
testing, elemental partitioning, and bulk analysis of sedimentary
samples.  Since each of these procedures measures a different property

of the sample, different storage requirements are required for samples

to be subjected to each testing procedure.  Therefore, this section

has also provided detailed guidance for the handling of sedimentary

samples from the time of collection until the time of analysis.
                                        C-17

-------
                              References
 1.  Environmental Protection Agency.   "Draft Interim Analytical
     Methods Manual for the Ocean Disposal Permit Program."
     EPA; Washington, D.C.  (January 197M-

 2.  Environmental Effects  Laboratory.   "Ecological Evaluation of
     Proposed Discharge of  Dredged or  Fill Material into Navigable
     Waters.  Interim Guidance for Implementation of Section UoMb)(l)
     of Public Law 92-500 (Federal Water Pollution Control Act
     Amendments of 1972)."   Miscellaneous Paper D-76-17.  U. S.
     Army Engineer Waterways Experiment Station, CE; Vicksburg,
     Mississippi (1976).
 3.  Environmental Protection Agency.   "NPDES Compliance Sampling
     Manual."  Enforcement  Division, Office of Water Enforcement,
     Compliance Branch, EPA; Washington, D.C.  139 P. (1977).
 U.  American Public Health Association.  Standard Methods for the
     Examination of Water and Wasjtewater Including Bottom Sediments
     and Sludges.  lUth Edition.  American Public Health Association,
     New York, New York. 1193 p. (1975).
 5.  Sly, P. G.  "Bottom Sediment Sampling."  Proc. 12th Conf. Gr.
     Lakes Res.  pp. 883-898  (1969).
 6.  Howmiller, R. P.  "A Comparison of the Effectiveness of Ekman
     and Ponar Grabs."  Trans. Amer. Fish. Soc. 100:560-561* (1971).
 7.  Wigley, R. L.  "Comparative Efficiencies of Van Veen and Smith-
     Mclntyre Grab Samplers as Revealed by Motion Pictures."  Ecology
     1*8:168-169 (1967).
 8.  Hudson, P. L.  "Quantitative Sampling with Three Benthic Dredges."
     Trans. Amer. Fish. Soc. 99:603-607 (1970).
 9.  Christie, N. D.  "Relationship Between Sediment Texture, Species
     Richness, and Volume of Sediment  Sampled by a Grab."  Marine
     Biology 30:89-96 (1975).
10.  Brannon, J. M., Plumb  Jr., R. H., and Smith, I.  "Long Term Release
     of Contaminants from Dredged Material."  Technical Report D-78-l*9.
     U. S. Army Engineer Waterways Experiment Station, CE, Vickstmrg,
     Mississippi.  66 p. (1978).
11.  Brannon, J. M., Engler, R. M., Rose, J. R., Hunt, P. G., and
     Smith, I.  "Selective Analytical Partitioning of Sediments to
     Evaluate Potential Mobility of Chemical Constituents During
     Dredging and Disposal Operations."  Technical Report D-76-7.
     U. S. Army Engineer Waterways Experiment Station, CE; Vicksburg,
     Mississippi.  90 p. (1976).
12.  Jenne, E. A., and Luoma, S. N.  "Forms of Trace Elements in Soils,
     Sediments, and Associated Waters:  An Overview of Their Determi-
     nation and Biological Availability."  U. S. Geological Survey;
     Menlo Park,  California.  Presentation at Biological Implications
                              C-18

-------
     of Metals in the Environment.   15th Life Sciences Symposium;
     Hanford, Connecticut.   71* P-  (September 29 - October 1,  1975).

13.  Chen, K. Y., Gupta, S.  K., Sycip,  A.  Z., Lu, J.  C.  S., Knesevic, M.,
     and Choi, W. W.  "Research Study on the Effect of Dispersion
     Settling, and Resediznentation on Migration of Chemical Constituents
     During Open Water Disposal of Dredged Materials."  Contract Report
     D-76-1.  U. S. Army Engineer Waterways Experiment Station, CE;
     Vicksburg, Mississippi.  221 p. (1976).
Ik.  Lee, G. F.  "Chemical  Aspects of Bioassay Techniques for
     Establishing Water Quality Criteria."  Water Res. 7:1525-15'*6
     (1973).
15.  Environmental Protection Agency.  "Methods for Chemical  Analysis
     of Water and Wastes."   National Environmental Research Center,
     EPA; Cincinnati, Ohio.   EPA-625/6-7U-003.  298 p. (197*0.
16.  Sherma, J.  "Manual of Analytical Quality Control for Pesticides
     and Related Compounds  in Human and Environmental Samples."
     Contract 68-02-1727.  EPA Health Effects Research Laboratory;
     Research Triangle Park, North Carolina.  EPA-600/1-76-017.
     Unnumbered  (1976).
17.  Environmental Protection Agency.  "Handbook for Analytical
     Quality Control in Water and Wastewater Laboratories."
     Environmental Monitoring and Support Laboratory, Environmental
     Research Center; Cincinnati, Ohio.  Draft Report (1978).

18.  Delfino, J. J.  "Quality Assurance in Water and Wastewater
     Analysis Laboratories."  Water and Sewage Works 12U(7):79-8U
     (1977).
19.  Anon.  "Federal Environmental Monitoring:  Will the Bubble Burst?"
     QA Report.  Environmental Science and Technology 12:126U-1269
     (1978).
20.  Keeley, J. W., and Engler, R. M.  "Discussion of Regulatory
     Criteria for Ocean Disposal of Dredged Materials:  Elutriate
     Test Rationale and Implementation Guidelines."  Miscellaneous
     Paper D-7U-1U.  U. S.  Army Engineer Waterways Experiment
     Station, CE; Vicksburg, Mississippi  (l97M.
21.  Lee, G. F., Piwoni, M. D., Lopez, J. M., Mariani, G. M.,
     Richardson, J. S., Homer, D. H., and Saleh, F.   "Research Study
     for  the Development of Dredged Material Disposal Criteria."
     Technical  Report D-75-1U.  U. S. Army  Engineer Waterways
     Experiment Station, CE; Vicksburg, Mississippi  (1975).
22.  Plumb, R.  H., Jr.   "A  Bioassay Dilution Technique to Assess the
     Significance  of Dredged Material Disposal."  Miscellaneous Paper
     D-76-6.  U. S. Army Engineer Waterways Experiment Station, CE;
     Vicksburg, Mississippi.   16 p.  (1976).
23.  Neff,  J. W.,  Foster, R.  S., and Slowey,  J.  F.   "Availability of
     Sediment-Adsorbed  Heavy  Metals  to  Benthos with Particular Emphasis
     on  Deposit Feeding Infauna."   Technical Report D-78-U2.  U.  S.
                                        C-19

-------
Army Engineer Waterways Experiment Station, CE; Vicksburg,
Mississippi.  286 p. (1978).
Lee, G. F. and Plumb, R. H. Jr.  "Literature Review on Research
Study for the Development of Dredged Material Disposal Criteria."
Contract Report D-7U-1.  U. S. Army Engineer Waterways Experiment
Station, CE; Vicksburg, Mississippi.  ll*5 p. (1971*).
                        C-20

-------
               APPENDIX D



TOTAL ACID DIGESTION METHOD FOR SEDIMENT



        (RANTALA  AND  LOR ING 1975)

-------
REAGENTS

     1.    Concentrated hydrofluoric,  hydrochloric, and nitric acids
          [ultrex grade (Baker)  or equivalent]

     2.    2.5  percent boric acid - dissolve 50 g [ultrex grade (Baker)
          or equivalent] boric acid in 2 L distilled-deionized water.

     3.    30 percent hydrogen peroxide [ultrex grade  (Baker) or equivalent],

PROCEDURE

     1.    If mercury is to be analyzed, a separate, 0.4-g (wet) aliquot
          is  digested and  analyzed  by the cold vapor  AA technique
          detailed in Appendix B.   (Note that the  sample weight must
          be doubled to achieve the 0.01-mg/kg dry weight detection
          limit  specified in Table 6 of Appendix A.)

     2.    Dry a representative portion of sediment  overnight at 60 C.
          Grind  to  100 mesh.

     3.    Weigh  200-mg  dried sample  into a decomposition  vessel.
          (Vessels  of various sizes may be obtained  from:   (a) Savillex
          Corp.,  5325 Hwy. 101, Minnetonka, MN, USA 55343; (b) H.K. Morri-
          son  and Sons Ltd., Mount Uniacke, N.S., Canada;  (c) Parr
          Instrument Company,  211-53rd St.,  Moline,  IL, USA  61265;
          (d)  Uniseal Decomposition Vessels Ltd., P.O.  Box  9463,  31 094
          Haifa, Israel  -  distributed in the USA by  Columbia Organic
          Chemicals Co., P.O. Box 1045, Camden, SC 29020.)   Add  0.5 mL
          cone.  HN03.

     4.    In a fume hood,  warm on a  hot plate at low  setting and open
          to the atmosphere until  the  brown N02  fumes  are  absent.
          To aid digestion, a few drops of 30 percent  H202  may be
          added.
                                 D-l

-------
5.   Cool to  room temperature.  Add 0.75 mL cone.  HC1  and  3.0  ml
     cone.  HF.  Seal  the digestion vessel.

6.   Heat in an oven  for 2  h at the digestion temperature recommended
     by the manufacturer (105°C - 130°C).   (Note:   a longer digestion
     time may be necessary for some sediments if visible particulates
     are present after  the  addition of boric acid.  Some fluorides,
     such as  CaF2 are  insoluble in the digestion acids  and  appear
     as a greyish sludge.  These  will dissolve upon  addition
     of boric acid.   Also, flakes of black,  undigested carbon-
     aceous matter may  also be present.  These will  not dissolve
     upon boric acid addition,  but will  not  interfere with the
     analysis.)

7.   Remove vessel from oven  and  cool to room temperature.  Add
     20 mL of  2.5 percent boric acid.   Reseal vessel  and return
     to oven at recommended temperature for 1 h.

8.   Remove vessel from oven and cool to room temperature.  Quanti-
     tatively transfer  contents to a 25 ml polyethylene  volumetric
     flask and bring  to the mark with distilled-deionized water.

9.   Digestate may be analyzed  directly  using the recommended
     atomic spectroscopic instrumental techniques.
                             D-2

-------
                    MULTI-ELEMENT ANALYSIS OF SILICATE ROCKS AND  MARINE
                    SEDIMENTS BY ATOMIC ABSORPTION SPECTROPHOTOMETRY

                                     R. T. T. Remain and D. H. Loring
                                      Department of ike Environment
                                       Fisheries and Marine Service
                                        Marine Ecology Laboratory
                                    Bedford Institute of Oceanography
                                     Dartmouth, Nova Scotia, Canada
ABSTRACT
A rapid method it described for the de-
tenninition of 21 dementi in silicate
tocki and marine sediments by flame
atomic •bwrpiion. The  method often
good accuracy and precision in major,
minor, and trace element determinations
aad it very suitable for marine f eochea-
ical studies. Precision and accuracy bare
been evaluated  by  analysing seven
TJ.S.G.S. rock standards.
RESUMC
On decril une meihode rapide pour la
determination de 21 elements dans des
roches silicates el de* sediments marins,
par absorption atomique de flsmme.
La methode donne une bonne exactitude
et une bonne precision pour la determi-
nation d'elementi majeurs, mineurs el
traces « esl Ires sdapfee  a des eludes
de feochinic marine. L'fvaluation de la
precision et de I'eianitude • M  faite
par 1'analyse de sept standards de rocbe
uses
ZUSAMMENFASSUNG
Es wird cine rascne Meihode fur Benin-
muttf vn 21 Qeaemea in Silikaiiestei-
nan  and  Meeres-Sedimenten mitielt
Fknunen-AtomsbsorptioD beacbrieben
Die Methode bietet fur die Bestimmunf
der Haupt.-Neben.-vnd Spurenelemente
cine jute Riehtifkeii und Primion und
tigntl aicn susieieichBet fur  leochem-
Ische Siudien  in Bereich des Meeres
Praasion und Richliikeit wurden durcb
Analyse  von 7 USCS Cesieins-Sid. be
slimmt.
INTRODUCTION

  Recent trends in the field of atomic absorption spectre*-
copy  have been toward flameless alemisation.  Severe
matrix problems are encountered, however, in samples such
as silicate rock* and marine sediments. Remedies such as
selective volatilization and the method of standard addi-
tions have failed to eliminate these problems in complex
solutions (1).
  Flame alomisation, for which matrix interferences can
generally  be controlled satisfactorily,  is one of the  most
suitable methods for (he analysis of silicate  rocks  and
marine sediment*.
  The method described here consists of an acid decom-
position in  a Teflon bomb  followed by  flame  atomic
absorption for the delerminstion of 21 elements in silicate
rocks and marine sediments.  It is more comprehensive in
terms of number of elements  than other reported schemes
based on HF decomposition (23,4).
  For the assessment of relative accuracy and the precision
of this method, U.S.G.S. rock standards granite G>2 and
diabase W-l have been analysed for 21 major, minor, and
trace elements: Al, Ca, Fe, K, Mg, Mn. Na, Si, Ti, Ba, Be,
Co. Cr, Cu. Li, Ni, Pb, Rb, Sr. V and Zn.
  In addition, U.S.C.S.  rock standards andesite AGV-1,
basalt BCR-1, dunile DTS-1, pranodiorile GSP-1 and peri-
                   dotite PCC-1 were analysed for 12 trace elements.
                     For some elements, notably Ba, Be, G, Pb, Rb and V.
                   which were determined by this method, no atomic absorp-
                   tion values were reported by Flanagan in his compilation
                   of the rock standard data (5).


                   PROCEDURE

                   Sample Preparation
                     Table 1 shows the analytical scheme for the  major and
                   trace elements.
                     For the major and some minor elements, a 0.1-g sample
                   is decomposed with HF and aqua regia in a Teflon vessel",
                   (6) and initially diluted to 100 ml. From this solution Mn.
                   Si, and also Ni for DTS-1 and  PCC-1 are determined
                   directly.
                     Solid KC1 is added to separate aliquots to give a potas-
                   sium concentration of 1500 ppm, ana these are anal) zed
                   forBa,Sr,andTi.
                     For the determinations of AX Ca, Fe, Mg,  and Na. a
                   further 20X dilution is made to give the desired dilution
                   factor of 2X10* for these elements in the KC1 (1500 pj.m)
                   matrix. A similar dilution is also made for the delrrinina-
                   tion of K by substituting NaCl as the diluent
                     For trace element determinations, a separate 1-p sample
                                                   TABIEI
                            Sample Preparation Diagram for Silicate Rock Analyses
   0.1-a, sample -+ Teflon vessel decomposition
                 -»1000X dilution
                 -»1000X dilution in 1500 ppm K
                 -»2000X dilution in 1500 ppm K
                 -»2000X dilution in 1500 ppm Na
   1-g sample —» Teflon vessel decomposition—» 100X dilution
             Mn,Si
             Ba. Sr, Ti
             Al, Ca. Fe, Mg, Na
             K

             Be, Co, Cr. Cu, li,
             Ni,Pb.Rb,V,Zn
                                                                  from H. X. Monijcn ind Sons Ltd. Mount llniicke,
                                                          H«nU Co., N. S., Cmnids.
ATOMIC  ABSORPTION NEWSUHER
Vol. 14. No. 5, September-October 1975
                                            D-3

-------
I-  r|rmmpo«rrl in ihr Trflon tr»»c! and ililulrrl  In 100 ml
Frutn lni« ."ululion Br, Co. Cr. Cu. Li. Ni. Pli. Rb. V. and
Zn are determined direct!).

Decomposition Proctdur*

   A finely ground sample of  0.1 ft or 1 g is weighed ac-
curately in the Teflon vessel and welled 1 ml of aqua regia.
Six ml of concentrated HF are added ilowl), and the cover
it  placed on liphlly. The vessel is immersed in a water bath
<90°C-100°C) for'l hr Care should be taken that the vessels
remain immersed during the entire heating period. Follow-
ing the heating, the vessels should immediate!) be removed
from the water bath and placed in cold  water. The water
level should not reach  the threaded cover. These precau-
tions are necessary to avoid sucking water into the vessels
by the vacuum created during cooling.  After cooling, the
contents are washed into a 100-ml  polypropylene volu-
metric flask containing 5.6 g H9BO, that has been shaken
with approximately 20 ml of  H90. The flask is shaken to
complete  the dissolution and  made  up to volume. The
aolutions are stored in polypropylene bottles.
   Sample solutions thus prepared are extremely stable. In
more concentrated trace element solutions,  a  gelatinous
precipitate of borosilicates will  form  due to the high con-
centration of silicon. This has no adverse effect on trace
element determinations  (6), and the removal of the bulk
of silicates is desirable.

Sf ondoref Preparation
   Single-element standards for major and minor elements
can be used. For the sake of convenience, however, a com-
bined slock standard solution  was prepared. This, as well
as other standards, was prepared from 1000-pprn standards.
The  1000-ppm standards were prepared either  in  the
laboratory or obtained commercially. No specific details
are given here for preparing these solutions. Such informa-
tion is readily available (7).
   The exceptions are the Si, Ti, and Cr standards which
are described below. A combined slock  standard solution
of 300 ug/ml Al and 100 ug/ml Ca. Fe, K, Mg, Mn, and
Na was prepared. This stock standard was diluted 25, 50,
75. 100. 200. and 400 times  for working standards. All
working standards were prepared in  1500 ppm K except
standards  for K which  were prepared in 1500 ppm Na.
The  decomposition blank contained 1 ml aqua regia. 6 ml
concentrated HF. and 5.6 g H3B03/100inl.
   A 1000-ppm Si standard was prepared from a silica rod
by the same decomposition procedure as described for the
samples. Working standards of 200-400 ppm  Si were pre-
pared in •  decomposition blank solution.
   A  1000-ppm Si siandard was prepared from a silica rod
bj  the »ame decomposition procedure as described for the
sample*. Working standards of 200-400 ppm  Si were pre-
pared in a  decomposition blank solution.
  The 1000-ppm Ti siandard was prepared from TiOj by
the same decomposition  procedure described  for the sam-
ples. Working standards of 2-10 ppm Ti were prepared in
the decomposition blank with 1500 ppm K.
   Single-element standards were prepared for Ba. Be, Cr,
Rb.  Sr. and V.  A combined  Mock siandard solution  of
50 tig -'ml each was prepared for Co, Cu. Li, Ni. Pb, and Zn.
V orkinp standards of 0.1-2.0 ppm for these elements were
prepared in the  decomposition  blank and are  stable for
»r\eral months.
   Working ftandards of 0.1-2.0 pfim Ba were prepared in
llir decomposition blank with 1500 ppm K.
   In preparing IV Mamiards  it  i* aHv^aMr to u*r  rom
 mernalh  atailaMr  stork «tjnr)arrl •><>lulion<> >inrr bcr\l
 Hum dust i» potential!)  loxic M orkinp Mandsrd* of 5-10
 ppb Be were prepared in ihr flrcompo*il>un blank
   A 1000-ppm Cr Manriard  wa*  prepared from K.-CrjO;
 Working standard* of 0 1-1.5 ppm Cr *ere prepared in the
 decomposition blank  The best semilixil) for Cr is obtained
 in the air-acetylene flame, but it is first  necessary to reduce
 Cr** to Cr14 in the solution because the standards give a
 low absorption if the Cr is not reduced. This reduction was
 carried out b) adding 2 ml perchloric  acid and 1 ml  307r
 H-O. to 20 ml of 1000 ppm Cr**, followed by healing and
 dilution to 1000 ml to obtain a 20-ppm  Cr** siandard (8).
 This standard waa diluted with the decomposition blank
 solution for working standards. Reduction of Cr** to Cr**
 is not necessary for Cr  determination  in a nitrous oxide-
 acetylene flame, but this method is less sensitive.
   Working standards of 0.2-3.0 ppm Rb were prepared in
 the decomposition blank with sufficient K added to match
 the sample concentration. The absorption of Rb  is sup-
  Eressed by the decomposition blank solution and enhanced
  ) K. Enhancement by K b progressive to about 600 ppm
 K, above which suppression by K occurs. Thus, the addition
 of excessive amounts of K results in a vastly suppressed
 absorption signal and cannot be recommended. Enhance-
 ment by  Na is small.  When  200 ppm  of Na  and K  were
 added to a 1-ppm Rb standard in tie decomposition blank.
 the combined  enhancement was about 2% more than  in
 200 ppm K only. Since the determination is highly depend-
 ent on K concentration, it was first necessary to determine
 the K content of the samples and match it to the standards.
 Since the  absorption of Rb was  observed to be linear at
 least to 3 ppm Rb for a particular K concentration, several
 standard curves could be drawn by using just one standard
 in each case. Where large numbers of samples are analyxed.
 it is not practical to match the sample  concentration  of K
 in each standard, but the Rb concentration can be inter-
 polated from other standard curves. It remains to be in-
 vestigated whether another alkali metal could be used  in
 large amounts to compensate for interferences without
 causing suppression of Rb absorption.
   Working standards of 0.1-1.0 ppm Sr were prepared in
 the decomposition blank with 1500 ppm K.
   Working standards of 0.3-5.0 ppm V were prepared  in
 the decomposition blank with 100 ppm  Al and 500 ppm K.
 Both Al  and  K enhance the  vanadium  absorption. No
 further enhancement was observed above 100 ppm Al and
 500 ppm K.
APPARATUS

  A Perkin-Elmer Model 303 atomic absorption speclro-
photometer equipped with a Deuterium Background Cor-
rector and a 10-mV full-scale deflection recorder was used
for the determination of all elements except Pb which was
determined using a Perkin-Elmer Model 306 atomic  ab-
sorption spectropholomeler. Scale expansion was 10X or
less for all determinations. Light sources for K, Na, and
Rb were Osram arc-discharge lamps. A Perkin-Elmer EDL
was  the light source for Pb. The Be light  source was a
Westinghouse  hollow  cathode lamp, and the  remaining
elements were determined with Perkin-Elmer Inlensitron*'
single-element  hollow  cathode lamps.  Acetylene was  thr
fuel for all elements. Air was used as the oxidant, except
for Al. Ca, .Mg, Si. Ti. Ba. Be, Sr, and V for  which nitrous
oxide was emplo)ed.
                                                     0-4
               ATOMIC  ABSORPTION NEWSLETTER
               Vol. 14, No. 5. September-October  1975

-------
 INTERFERENCES

   Matrix mterfrrrnrr* on major and minor element* are
 reduced b\ dilution in formp the concenlration to the linear
 portion of the ar»*orhance curve, loniution is controlled b>
 additions of 1500 ppm K or Na to standard." and samples.
 Chemical suppression  by elements such as silicon  and
 aluminum  on calcium  and  magnesium absorption  was
 avoided b\ the use of a nitrous oxide-acel)lene flame. This
 eliminated the need for lanthanum which is essential in an
 air-acetylene flame. No interferences  were observed for
 Mn and Si in a fluobonc-boric acid matrix.
   lonization in Ba and Sr determinations  Has controlled
 by adding  1SOO ppm  K  to samples and standards. Stan-
 dards for Cr, Rb,  and V require special attention as dis-
 cussed under standard preparation. Molecular absorption
 is corrected by the Deuterium Background Corrector in
 Co, Ni, and Pb determinations. No interferences were ob-
 served for  Be, Cu, Li, and Zn  in  a fluoboric-boric acid
 matrix.

 RESULTS AND DISCUSSION

   Table 11  indicates the relative accuracy of the major and
 minor element data for W-l and C-2 by the comparison of
 our values  to those compiled by Flanagan (9). The preci-
 sion is expressed as a coefficient of variation. To obtain
 these data, 6 separate portions of W-l  and G-2  were
 analysed.
   Table  III indicates the relative  accuracy of the trace
 element determinations for teven U.S.G.S. standard rocks
 by comparison of our values to those compiled by Flanagan.
 The  precision is  expressed as a coefficient of variation
 (Table IV). These data are based on 3 separate portions

                       TABLE II
    Major Elements in U.S.G.S. Standard Rocks (%)

Al
Co
Fe
K
M9
Mn
Na
Si
Ti

W-l
G-2
W-l
G-2
W-l
G-2
W-l
G-2
W-l
G-2
W-l
G-2
W-l
G-2
W-l
G-2
W-l
G-2
A
7.94
8.15
7.83
1.39
7.76
1.85
0.53
374
3.99
0.46
0.132
0.026
1.59
3.02
24.6
32.3
0.64
0.30
B
7.68
8.11
7.83
MO
7.96
1.90
0.56
3.62
3.86
0.44
0.127
0.025
1.60
2.94
24.6
31.5
0.67
0.31
Coeff.
1.5
2.7
0.7
1.9
0.4
5.0
4.7
1.2
0.1
4.1
1.1
4A
1.2
0.8
2.6
0.8
1.5
2.6
A Flanagan (8)
B This work
nf thr rock -ample
   Clo*e a^rn-nirnt with Flanagan'* \alur- ran 1«- nolc.l Im
most clement* Zinc talues for DTS-1 ind PCC-1 arr )<•«
Tin* could be due to incomplete decomposition n( rhru
milrv An appreciable amount of chromite which rmilil not
br dcroinjio-rd entirel) is contained in DTS-1 and PCC-1 .
therefore, it was not possible to determine Cr accurnirU
in thene samples.
   Nickel values  for  ACV-1  and W-l are comparatively
low. An attempt  was made to determine Ni without back-
ground correction. The positive error that resulted, how-
ever, was considerably larger than the negative error w ith
background correction. It has been shown that very large
positive or negative error* are possible  in the determina-
tion of Pb, Zn, Ni, and Co in geological materials which
depend upon the concentration of the trace elements and
major elements in  the sample solution (101. The dilution
factor of 100X and the fluoboric-boric acid matrix used in
this method minimize these errors.
  The precision  for major and minor elements is  ^ 55
and for trace elements generally < 10%.
  For most  major elements the relative accuracy is com-
parable and  precision  improved  over  that  reported b>
Buckley and Cranston. However, for trace elements present
in low concentrations, both the relative accuracy and pre-
cision are significantly improved over the values reported
by the above authors. No complete  comparison can be
made  at this time  with other scheme; due to the lack of
published statistical  data  on all  the elements we hate
studied by this method.
  The method described here is ideal for major geochemi-
cal studies on marine sediments (11) and is now used in
this laboratory for routine analyses. Although the organic
matter is not completely oxidized during the decomposi-
tion, the residue is  generally  <3#, and  its elemental
contribution is neligible. The method is rapid as a larpe
number of inexpensive Teflon vessels can be used simul-
taneously and the  dilutions carried out with commercial
sampler-diluters.

CONCLUSION

  The  method described  for  multi-element analyse.-  of
silicate rocks and marine sediments by flame atomic ab-
                                                                                TABLE IV
                                                                 Precision (Coefficient of VoriotionJ for the
                                                                    Determination of Trace Elements in
                                                                      * U.S.G.S. Standard Rockt(% |

Bo
Be
Co
Cr
Cu
Li
Ni
Pb
Rb
Sr
V
Zn
AGV-1
2.0
2.8
14
5.6
4.2
2.6
—
4.3
2.3
2.2
1.2
1.3
BOM
5.7
2.2
7.3
3.9
6.7
2.5
—
15
1.2
2.9
2.4
1.6
W-l
3.2
4.0
2.4
0.5
3.7
2.2
4.5
—
4.1
5.1
0.7
0.7
ors-t
—
_
0.9
_
4.2
6.4
—
—
—
—
—
1.5
G-2
1.1
1.7
7.9
6.1
4.3
0.8
—
11
2.1
1.6
6.0
4.1
GSP-1
1.4
1.9
12
1.9
4.9
1.5
—
8.6
0.5
3.5
4.3
10
PCC-I
_
_
09
_
5.6
10
—
—
_
—
7.5
18
ATOMIC  ABSORPTION NEWSLETTER
Vol. 14. No. 5. September-October 1975
                                              D-5

-------
                                                    TABLE 111
                                 Trace Elements in U.S G S. Standard Rockt I pom)

la (a)
(b)
»• (e)

(b)
Pb{o)
(b)
Rb(o)
(b)
Sr (a)
(b)
V (a)
(b)
Zn(a)
(b)
a Flanagan (8)
b Thitwork
AGV-J
1208
1140
3
2.6
14
18
12
9
40
59
12
12
18.5
12
35
40
67
67
657
698
125
128
84
90


ICft-7
675
630
1.7
2.1
38
36
18
11
18
15
13
14
16
<12
16
16
47
49
330
363
399
446
120
129


W-1
160
156
0.8
0.9
47
41
114
116
110
122
14
14
76
66
8
<«
21
21
190
194
264
285
86
86


075- r
2.4
<5
_
<0.5
133
128
—
not analyzed
7
7
2
2.7
2269
2420
14
19
_
not analyzed
0.35
<2
10
<15
45
32


G-2
1870
1970
2.6
2.9
5.5
7
7
7
12
12
37
34
5
<12
31
34
168
166
479
457
35
42
85
87


GSM
1300
1230
13
1.7
6
9
12
12
33
31
32
30
12.5
<12
51
59
254
248
233
245
53
62
98
105


PCC-I
1.2
<5
—
.
 B. F. J. Jolinmn. T. C  Woodi*. Jr. and J. M. Cumminc», Jr.. Au
   Absorption Ne»»lcH- II. 118 (1972).
 9. F. J. Flintpjo, Ceaeblm. Cotmotb'm. Aeu J7,1189 (1973).
 10. C. J. S. Covett and R. E.  Vhiiebecd. J. Ceeehem. E«pl t. 121
   11973).
 II. D. H. Urine tnd D. J. C. Noia. Bulletin 182, Fiskeriw Resetrcb
   Board of Canida.
                                                           D-6
                ATOMIC  ABSORPTION NEWSLETTER
                Vol. H, No. S, September-October  1975

-------
              APPENDIX E



HN03/HC104 DIGESTION METHOD FOR TISSUE



          (TETRA TECH 1986a)

-------
11.0  PROCEDURE

11.1  Homogenize  samples prior to analysis  to  ensure that a representative
aliquot is taken.  Any  grinder or homogenizer that has  been found to  be
free of  contamination may  be used.  Samples  should be ground wet to avoid
losses of volatile elements  (e.g., Hg, Se) during drying.  The liquid associated
with a sample after  thawing should be retained  as  part of the sample.

11.2  Transfer  the  sample paste  to a container  suitable for storage.   If
not immediately  analyzed, the  samples should  be  frozen (-20° C)  until required.
Containers should  be tightly sealed to prevent  moisture loss or gain during
storage.

11.3  Dry Weight  Determination - if sample size permits and dry-weight
concentrations are required, dry  weight  determinations  may  be  performed
as follows:   transfer an  aliquot of approximately  3 g (weighed to the  nearest
0.1 g) to a pre-weighed  dish.  Allow the  sample to dry in an oven at  105° C
overnight, and determine  the  solid  residue weight  to  the nearest  0.1  g.
The percent total  solids  is calculated as:

          TS = [dry  residue wt  (g)]/[west  sample wt (g)]

Dry-weight determinations should  not  be made  at  the cost of having insufficient
sample for metals  analysis.   Significant decreases in the size of samples
used for  extraction  will  decrease  attainable  detection limits.

11.4  Accurately  weigh  representative aliquots  of homogenized tissue to
the nearest 0.1  mg.   If sample  size permits,  approximately 5 g is required
to maintain optimum detection limits.   Transfer the weighed tissue to a
pre-cleaned 125-mL Erlenmeyer  flask equipped  with  an all-glass reflux cap.
Analyze  a sufficient  number of  reagent  blanks,  sample duplicates, analyte
spikes, and certified reference materials  concurrently.
                                    E-l

-------
11.5  Add  10.0 ml of concentrated nitric acid (ACS grade or better), replace
cap and swirl.  Allow  flask  to stand at room  temperature for  about 15 h
in a dust-free  ventilated  environment.   Periodically swirl the contents
to help solubilize the tissue.

11.6  After 15 h, gently heat the  flask  to approximately  100° C - hold
at this temperature  for 1  h.   Gradually  increase the temperature in 50° C
increments  to  a  maximum of 250° C.  Continue digesting until  all  tissue
has been solubilized.  This usually takes about 4 h.  Do not rush  the initial
digestion  as losses of volatile elements will likely occur.   Once digestion
is complete, cool flasks to room temperature  and  add 4.0 ml  of  perchloric
acid.

     CAUTION:  Perchloric acid  is a strong  oxidizing agent. The analyst
              must be fully aware of the precautions  associated with
              its use.   This procedure (i.e., use of perchloric acid
              at sub-fuming temperatures)  has been  safety performerd
              without a perchloric hod, but a perchloric hod is strongly
              recommended nonetheless.   Laboratories  that do not
              carefully monitor  perchloric acid  digestions will be
              endangered without perchloric hoods.  Safety precautions
              and background information pertaining  to perchloric
              acid  can be found  in Schilt (1979; Perchloric Acid
              and Perchlorates.  G. Frederick Smith Chemical Company,
              Columbus, OH.).

11.7  Return flasks  to  the hotplate which has been cooled to about 200° C.
Continue  heating for 1 h, then increase plate temperature to 300° C.  Hold
at this temperture until all traces of nitric  acid fumes  have disappeared
and the solutions have become clear.   Do not overheat flasks or  allow perchloric
fumes (dense white) to appear.  If  perchloric fumes appear, reduce heat
immediately.  Remove the exracts when clear and cool them to room  temperature.
                                    E-2

-------
11.8  When the digestion is complete,  rinse the caps into the flasks  and
transfer the  extract to a pre-cleaned 100-mL volumetric  flask.   Rinse  the
Erlenmeyer flask  three times with DDW and  combine with the extract previously
added to the  volumetric  flask.  Adjust the volume with  DDW and transfer
to a precleaned plastic bottle.

      NOTE:  Some  elements are not  as  stable as others in solution  and
therefore should  be analyzed first.  Stability can be determined by daily
analysis of the extracts.  However, the following can be used as  a guideline:

          Sb,  Pb, Hg, Se, and Ag - analyze within 1 day
          As  and  Cd - analyze within 2 days
          Cr,  Cu, Ni and Zn - analyze within 1 wk
          Be  and  Tl - to be determined.
                                    E-3

-------
                APPENDIX F





APDC/MIBK EXTRACTION METHOD FOR SALT WATER






          (GREENBURG ET  AL.  1985)

-------
   303  B.  Determination of  Low Concentrations of Cadmium,
   Chromium, Cobalt, Copper, Iron, Lead, Manganese, Nickel,
    Silver, and  Zinc by Chelation with Ammonium Pyrrolidine
   Dithiocarbamate (APDC) and Extraction into Methyl Isobutyl
                           Ketone (MIBK)
1. Apparatus
  a. Atomic absorption spectrometer and as-
sociated equipment: See Section 303.2.
  b. Burner head, conventional. Consult
manufacturer's operating manual for sug-
gested burner head.


2.  Reagents
  a. Air See 303A.20.
  b. Acetylene: See 303A.24.
  c. Metal-free water: See 303A.2c.
  d.  Methyl uobutyl ketone (MIBK). re-
agent grade. For trace analysis,  purify
MIBK by redistillation or by sub-boiling
distillation.
  e. Ammonium pyrrolidine dithiocarba-
mate (APDC) solution: Dissolve 4 g APDC
in 100 mL water. If necessary, purify
APDC with an equal volume of  MIBK.
Shake 30 s in a separatory funnel,  let sep-
arate, and withdraw lower portion. Discard
MIBK layer.
  / Nitric acid. HNO,, cone, ultrapure.
  g. Standard metal solutions: See 303A.2/.
  h.  Potassium  permanganate solution,
KMnO,, 5% aqueous.
  i. Sodium sulfate. Na,SO4, anhydrous.
  / Water-saturated MIBK: Mix one part
                                 F-l

-------
ATOMIC ABSORPTION SPECTROMETRY
                                                           161
purified MIBK with one pan water in a
separatory funnel. Shake 30 s and let sep-
arate. Discard aqueous layer. Save MIBK
layer.
  k. Sodium azide solution: Dissolve 0.1 g
NaN, in water and dilute to  100 mL.

3. Procedure
  a. Instrument  operation:  See  Section
303A.3a. After final adjusting of burner
position, aspirate water-saturated MIBK
into flame and gradually reduce  fuel flow
until flame is similar to that before aspi-
ration of solvent.
  b. Standardization: Select at least three
standard metal solutions (prepared as in
303A.2/) to bracket expected  sample metal
concentration and to be, after extraction,
in the optimum concentration range of the
instrument. Adjust 100 mL of each stand-
ard and 100 mL of a metal-free water blank
to pH 3 by adding \N HNO, or \N NaOH.
For individual element extraction, use  the
following pH  ranges to  obtain  optimum
extraction efficiency:
   Element
pH Range for Optimum
     Extraction
     Ag          2-5 (complex unstable)
     Cd                  1-6
     Co                  2-10
     Cr                  3-9
     Cu                 0.1-8
     Fe                  2-5
     Mn         2-4 (complex unstable)
     Ni                  2-4
     Pb                 0.1-6
     Zn                  2-6


 NOTE For Ag and Pb extraction the op-
 timum pH value is 2.3  ±  0.2. The  Mn
 complex deteriorates  rapidly, resulting in
 decreased instrument response. Unless Mn
 can be analyzed immediately after extrac-
 tion, use another analytical  procedure.
  Transfer each  standard  solution  and
blank to  individual  200-mL  volumetric
flasks,  add  1  mL APDC  solution,  and
shake to mix. Add 10 mL MIBK and shake
vigorously for 30 s. (The maximum volume
ratio of sample to MIBK is 40.) Let con-
tents of each flask separate into aqueous
and  organic  layers,  then  carefully  add
water down the side of each flask to bring
the organic layer into the neck and acces-
sible to  the aspirating tube.
  Aspirate organic extracts directly  into
the flame (zeroing instrument on a water-
saturated  MIBK  blank)  and  record ab-
sorbance.  With some instruments it  may
be necessary to convert percent absorption
to absorbance by using  a table generally
provided by the manufacturer.
  Prepare a calibration curve by  plotting
on linear graph paper absorbances of ex-
tracted  standards against their concentra-
tions before extraction.
  c. Analysis of samples: Rinse atomizer by
aspirating  water-saturated  MIBK. Aspi-
rate  organic extracts treated as above di-
rectly   into   the  flame   and  record
absorbances.
  With the  above  extraction procedure
only hexavalent chromium is measured. To
determine  total chromium, oxidize triva-
lent chromium to hexavalent chromium by
bringing sample to a boil and  adding suf-
ficient KMnO. solution dropwise to give a
persistent pink color while the solution is
boiled for 10 min. Destroy excess KMnO.
by adding 1 to 2 drops NaN, solution  to
the boiling solution, allowing 2 min for the
reaction to proceed. If pink color persists,
add  1 to 2 more drops NaN, solution and
wait 2 min. Heat an additional 5 mm. Cool,
extract with MIBK, and aspirate.
   During extraction, if an emulsion forms
at the  water-MIBK interface, add anhy-
drous Na2SO, to obtain a homogeneous or-
ganic phase.
   Determine  recovery of silver whenever
 matrices change.
  162
                                                    METALS (300)
    To avoid problems associated with in-
  stability of extracted metal complexes, de-
  termine metals immediately after extrac-
  tion.
                          4. Calculations
                            Calculate the  concentration  of each
                          metal ion in micrograms per liter by re-
                          ferring to the appropriate calibration curve.
                                  F-2

-------
                   APPENDIX G









DFAA INSTRUMENTAL AND SPECTROPHOTOMETRIC METHODS






                (U.S. EPA 19793)

-------
                                    ANTIMONY

               Method 204.1  (Atomic absorption, direct aspiration)

                                                        STORE! NO. Total  01097
                                                                     Dissolved 01095
                                                                    Suspended 01096

Optimum Concentration Range:   1-40 mg/1 using a wavelength of 217.6 nm
Sensitivity:     0.5 mg/1
Detection Limit:     0.2 mg/1

Preparation of Standard Solution
     1.   Stock Solution: Carefully weigh 2.7426 g of antimony potassium tartrate (analytical
          reagent grade) and dissolve in deionized distilled water. Dilute to 1 liter with deionized
          distilled water. 1 ml = 1 mgSb (1000 mg/1).
     2.   Prepare dilutions of the stock solution to be used as calibration standards at the time of
          analysis. The calibration standards should be prepared using the same type of acid and at
          the same concentration as will result in the sample to be analyzed either directly or after
          processing.

Sample Preservation
     1.   For sample handling and preservation, see part 4.1 of the Atomic Absorption Methods
          section of this manual.

Sample Preparation
     1.   The procedures for preparation of the sample as given in parts 4.1.1 through 4.1.4 of the
          Atomic Absorption Methods section of this manual have been found to be satisfactory.

Instrumental Parameters (General)
     1.   Antimony hollow cathode lamp
     2.   Wavelength: 217.6 nm
     3.   Fuel: Acetylene
     4.   Oxidant: Air
     5.   Type of flame: Fuel lean

Analysis Procedure
     1.   For analysis procedure and calculation, see "Direct Aspiration", part 9.1 of the Atomic
          Absorption Methods section of this manual.


Approved for NPDES
Issued  1974
Editorial revision 1978

                                      G-l

-------
Interferences
     1.    In the presence of lead (1000 mg/1), a spectral interference may occur at the 217.6 nm
           resonance line. In this case the 231.1 nm antimony line should be used.
     2.    Increasing acid concentrations decrease antimony absorption. To avoid this effect, the
           acid concentration in the samples and in the standards should be matched.

Notes
     1.    Data to be entered into STORET must be reported as ug/I.
     2.    For concentrations of antimony below 0.35 mg/1, the furnace procedure (Method 204.2)
           is recommended.

Precision and Accuracy
     1.    In  a single  laboratory (EMSL), using a mixed  industrial-domestic waste effluent at
           concentrations  of 5.0  and IS mg Sb/1, the standard deviations were ±0.08 and  ±0.1,
           respectively. Recoveries at these levels were 96% and 97%, respectively.
                                                G-2

-------
                                     ARSENIC

              Method 206.3 (Atomic  Absorption—gaseous  hydride)
                                                        STORET NO.  Total 01002
                                                                    Dissolved 01000
                                                                   Suspended 01001
1.    Scope and Application
     1.1   The gaseous  hydride  method  determines  inorganic  arsenic  when  present  in
          concentrations at or above 2 ug/1. The method is applicable to drinking water and most
          fresh and saline  waters in the absence of high concentrations of chromium, cobalt,
          copper, mercury, molybdenum, nickel and silver.
2.    Summary of Method
     2.1   Arsenic in the sample is first reduced to the trivalcnt form using SnClj and converted to
          arsine, AsH3, using zinc metal.  The gaseous hydride is swept into an argon-hydrogen
          flame of an atomic absorption spectrophotometer. The working range of the mehtod is
          2-20 ug/1. The 193.7 nm wavelength line is used.
3.    Comments
     3.1   In analyzing drinking water and most surface and ground waters, interferences are rarely
          encountered. Industrial waste samples should be spiked with a known amount of arsenic
          to establish adequate recovery.
     3.2  Organic forms of arsenic must be converted to inorganic compounds and organic matter
          must  be  oxidized before beginning the analysis. The oxidation procedure given in
          Method 206.S (Standard Methods, 14th Edition, Method 404B, p. 285, Procedure 4.a)
          has been found suitable.
     3.3  For sample handling  and preservation, see part 4.1 of the Atomic Absorption Methods
          section of this manual.
     3.4  For quality control requirements and optional recommendations  for use in drinking
          water analyses, see part lOof the Atomic Absorption Methods section of this manual.
     3.5  Data to be entered into STORET must be reported as ug/1.
4.   Precision and Accuracy
     4.1  Ten replicate solutions of o-arsenilic acid at the 5,10 and 20 ug/1 level were analyzed by
          a single laboratory. Standard deviations were ±0.3, ±0.9 and ±1.1 with recoveries of 94,
          93 and  85%, respectively.  (Caldwell, J. S., Lishka,  R. J., and  McFarren, E.  F.,
           "Evaluation of a Low Cost Arsenic and Selenium Determination at Microgram per Liter
           Levels", JAWWA., vol 65, p 731, Nov., 1973.)
 Approved for NPDES and SDWA
 Issued  1974
                                         G-3

-------
5.    References

     3.1.   Except for the perchloric acid step, the procedure to be used for this determination is
           found in: Standard  Methods  for the Examination  of Water and  Wastewater, 14th
           Edition, p!59, Method 301 A(VII),( 1975)
                                             G-4

-------
                                      ARSENIC

                     Method 206.4  (Spectrophotometric-SDDC)

                                                                STORE! NO.  01002
                                                         Inorganic,  Dissolved  00095
                                                              Inorganic, Total  00997
                                                        Inorganic,  Suspended  00996

1.    Scope and Application
     1.1  The silver diethyldithiocarbamate method determines inorganic arsenic when present in
          concentrations at or above 10 ug/1. The method is applicable to drinking water and most
          fresh  and saline waters  in the absence of. high concentrations of chromium, cobalt,
          copper, mercury, molybdenum, nickel, and silver. Domestic and industrial wastes may
          also be analyzed after digestion (see 3.3).
     1.2  Difficulties may be encountered with certain industrial  waste  materials containing
          volatile substances. High sulfur content of wastes may exceed removal capacity of the
          lead acetate scrubber.
2.    Summary of Method
     2.1  Arsenic in the sample is reduced to arsine, AsH3,  in acid solution in  a hydrogen
          generator. The arsine is passed through a scrubber to remove sulfide and is absorbed in a
          solution of silver diethyldithiocarbamate  dissolved in pyridine. The red complex thus
          formed is measured in a spectrophotometer at 535 nm.
3.    Comments
     3.1  In analyzing drinking water and most surface and ground waters, interferences are rarely
          encountered. Industrial waste samples should be spiked with a known amount of arsenic
          to establish adequate recovery.
     3.2  It is essential that the system be airtight during evolution of the arsine, to avoid losses.
     3.3  If concentration of the sample and/or oxidation of any organic matter is required, refer
          to Method 206.5. [Standard Methods. 14th Edition, Method 404B, p. 284, Procedure 4.a
          (1975)]. For sample handling and preservation, see part 4.1 of the Atomic Absorption
          Methods section of this manual.
          3.3.1  Since nitric acid gives  a negative interference in this test, use sulfuric acid as a
                preservative if only inorganic arsenic is being measured.
     3.4   1-Ephedrine  in chloroform  has been  found  to  be  a  suitable solvent  for  silver
          diethyldithiocarbamate if the analyst finds the  odor of pyridine objectionable [Anal.
          Chem. 45,1786(1973)].
     3.5  For quality control requirements and optional  recommendations for use in drinking
          water analyses, see part 10 of the Atomic Absorption Methods section of this manual.


Approved for  NPDES and SDWA
Issued 1971
Editorial revision 1974
                                        G-5

-------
4.    Precision and Accuracy
     4.1  In a round-robin study reported by Standard Methods a synthetic unknown sample
          containing 40 ug/1, as As, with other metals was analyzed in 46 laboratories. Relative
          standard deviation was ±13.8% and relative error was 0%.
5.    Reference
     5.1  The procedure to be used for this determination is found in:
          Standard Methods for the Examination of Water and Wastewater, 14th Edition, p. 283,
          Method 404A( 197 5).
                                               G-6

-------
                                     ARSENIC

           Method  206.5   (Sample Digestion Prior to Total Arsenic
             Analysis by Silver Diethyldithiocarbamate or Hydride
                                     Procedures)

1.    Scope and Application:
     1.1   Both the silver diethyldithiocarbamate spectrophotometric method and the AA hydride
          procedure measure inorganic arsenic. Therefore, if either of these procedures are being
          employed for the purpose of measuring total arsenic (inorganic plus organic), all
          organically  bound arsenic must first be converted to an inorganic form prior to the
          analytical determination. This may be accomplished with H]S04-HNOj.
2.    Procedure
     2.1   To a suitable sample containing from 2 to 30 ug of arsenic, add 7 ml (1 +1) H2SO4 and 5
          ml cone HNO3. Evaporate the sample to SO3 fumes. Caution:   If the sample chars, stop
          the digestion immediately, cool and add additional cone HNO3. Continue digestion
          adding additional cone HNO, as necessary.
     2.2   If the sample remains colorless, or straw-yellow  during evolution of SO3 fumes, the
          digestion is complete.
     2.3   Cool the digested sample, add about 25 ml distilled water, and again evaporate to SO}
          fumes to expel oxides of nitrogen.
     2.4   The sample is now ready for analysis using either the hydride or spectrophotometric
          procedure.
3.    Interferences
     3.1   All traces of nitric acid must be removed before either the spectrophotometric or the
          hydride procedures are applied. Oxides of nitrogen should be expelled by taking the
          sample to fumes of SO3.
4.    Notes
     4.1   The digestion step may be carried out in a flask on a hot-plate or in a Kjeldahl apparatus.
          This digestion step may also be used, in effect, to concentrate the sample, inasmuch  as
          any size volume may be processed.

                                      Bibliography

 1.    Standard Methods for the Examination of Water and Wastewater, p285, method 404B, step 4a,
      14th Edition (1975).
 Approved  for NPDES and SDWA
 Issued  1978
                                          G-7

-------
                                    CADMIUM

              Method  213.1 (Atomic  Absorption, direct aspiration)

                                                        STORE! NO.  Total  01027
                                                                     Dissolved  01025
                                                                   Suspended  01026

Optimum Concentration Range:   0.05-2 mg/1 using a wavelength of 228.8 nm
Sensitivity:     0.025 mg/1
Detection Limit:      0.005 ing/1

Preparation of Standard Solution
     1.   Stock Solution: Carefully weigh 2.282 g of cadmium sulfate (3CdSO«»8H,O, analytical
          reagent grade) and dissolve in deionized distilled water. 1 ml =  1  mg Cd (1000 mg/1).
     2.   Prepare dilutions of the stock solution to be used as calibration standards at the time of
          analysis. The calibration standards should be prepared using the same type of acid and at
          the same concentration as will result in the sample to be analyzed either directly or after
          processing.

Sample Preservation
     1.   For sample handling and preservation, see part 4.1 of the Atomic Absorption Methods
          section of this manual.

Sample Preparation
     1.   The procedures for preparation of the sample as given in parts 4.1.1 through 4.1.4 of the
          Atomic Absorption Methods section of this manual have been found to be satisfactory.

Instrumental Parameters (General)
     1.   Cadmium hollow cathode lamp
     2.   Wavelength: 228.8 nm
     3.   Fuel: Acetylene
     4.   Oxidant: Air
     5.   Type of flame: Oxidizing

Analysis Procedure
      1.    For analysis procedure and calculation, see "Direct Aspiration", part 9.1 of the Atomic
           Absorption Methods section of this manual.
 Approved  for NPDES and SDWA
 Issued  1971
 Editorial revision  1974
                                        G-8

-------
Notes
     1.    For levels of cadmium below 20 ug/1, either the Special Extraction Procedure given in
           Part 9.2 of the Atomic Absorption methods section as the furnace technique, Method
           213.2 is recommended.
     2.    Data to be entered into STORET must be reported as,ug/l.
     3.    For quality control requirements and optional  recommendations  for use in drinking
           water analyses, see part 10 of the Atomic Absorption Methods section of this manual.

Precision and Accuracy
     1.    An interlaboratory study on trace metal analyses by atomic absorption was conducted by
           the Quality Assurance and Laboratory  Evaluation Branch of EMSL.  Six synthetic
           concentrates containing varying levels of aluminum, cadmium, chromium, copper, iron,
           manganese, lead and zinc were added to natural water samples. The statistical results for
           cadmium were as follows:

                                                         Standard
Number        True Values          Mean Value           Deviation           Accuracy as
of Labs           ug/liter              ug/Iiter              ug/liter               % Bias

  74              71                   70                  21                  -2.2
  73              78                   74                  18                  -5.7
  63              14                   16.8                 11.0                  19.8
  68              18                   18.3                 10.3                   1.9
  55                1.4                  3.3                  5.0                 135
  51                2.8                  2.9                  2.8                   4.7
                                              G-9

-------
                                      COPPER
              Method  220.1 (Atomic Absorption, direct  aspiration)

                                                        STORET  NO. Total 01042
                                                                     Dissolved 01040
                                                                    Suspended 01041

Optimum Concentration Range:  0.2-5 mg/1 using a wavelength of 324.7 nm
Sensitivity:     0.1 mg/1
Detection Limit:      0.02 mg/1

Preparation of Standard Solution
      1.    Stock Solution: Carefully weigh 1.00 g of electrolyte copper (analytical reagent grade).
           Dissolve in 5 ml redistilled HNO, and make up to 1 liter with deionized distilled water.
           Final concentration is 1 mg Cu per ml (1000 mg/1).
      2.    Prepare dilutions of the stock solution to be used as calibration standards at the time of
           analysis. The calibration standards should be prepared using the same type of acid and at
           the same concentration as will result in the sample to be analyzed either directly or after
           processing.

Sample Preservation
      1.    For sample handling and preservation, see part 4.1 of the Atomic Absorption Methods
           section of this manual.

Sample Preparation
      1.    The procedures for preparation of the sample as given in parts 4.1.1 thru 4.1.4 of the
           Atomic Absorption Methods section of this manual have been found to be satisfactory.

 Instrumental Parameters (General)
      1.    Copper hollow cathode lamp
      2.    Wavelength: 324.7 nm
      3.    Fuel: Acetylene
      4.    Oxidant: Air
      5.    Type of flame: Oxidizing

 Analysis Procedure
      1.    For analysis procedure and calculation, see "Direct Aspiration", part 9.1 of the Atomic
           Absorption Methods section of this manual.
 Approved for NPDES
 Issued 1971
 Editorial revision 1974 and  1978
                                        G-10

-------
Notes
      1.    For levels of copper below SO i/g/1, either the Special Extraction Procedure, given in part
           9.2 of the Atomic Absorption Methods section or the furnace technique. Method 220.2,
           is recommended.
     2.    Numerous absorption lines are available for the determination of copper.  By selecting a
           suitable absorption wavelength, copper samples may be analyzed over a very wide range
           of concentration. The following lines may be used:
           327.4 nm Relative Sensitivity 2
           216.S nm Relative Sensitivity 7
           222.S nm Relative Sensitivity 20
     3.    Data to be entered into STORET must be reported as ug/1.
     4.    The 2,9-dimethyl-1, 10-phenanthroline colorimetric method may also be used (Standard
           Methods, 14th Edition, p. 196).

Precision and Accuracy
      1.    An interlaboratory study on trace metal analyses by atomic absorption was conducted by
           the  Quality Assurance and Laboratory Evaluation  Branch of EMSL.  Six  synthetic
           concentrates containing varying levels of aluminum, cadmium, chromium, copper, iron,
           manganese, lead and zinc were added to natural water samples. The statistical results for
           copper were as follows:
                                                          Standard
 Number        True Values          Mean Value           Deviation           Accuracy as
 of Labs          ug/liter              ug/liter               tig/liter              % Bias

  91              302                 305                  56                    0.9
  92              332                 324                  56                   -2.4
  86               60                  64                  23                    7.0
  84               75                  76                  22                    1.3
  66                7.5                  9.7                 6.1                 29.7
  66               12.0                 13.9                 9.7                 15.5
                                           G-ll

-------
                                        IRON
              Method 236.1  (Atomic Absorption,  direct aspiration)

                                                        STORE! NO.  Total 01045
                                                                     Dissolved 01046
                                                                    Suspended 01044

Optimum Concentration Range:   0.3-5 mg/1 using a wavelength of 248.3 nm
Sensitivity:     0.12mg/l
Detection Limit:       0.03 mg/1

Preparation of Standard Solution
     1.    Stock Solution: Carefully weigh 1.000 g of pure iron wire (analytical reagent grade) and
           dissolve in 5 ml redistilled HNO3, wanning if necessary. When solution is complete make
           up to 1 liter with deionized distilled water. 1 ml = 1 mg Fe (1000 mg/1).
     2.    Prepare dilutions of the stock solution to be used as calibration  standards at the time of
           analysis. The calibration standards should be prepared using the same type of acid and at
           the same concentration as will result in the sample to be analyzed either directly or after
           processing.

Sample Preservation
     1.    For sample handling and preservation, see part 4.1 of the Atomic Absorption Methods
           section of this manual.

Sample Preparation
     1.    The procedures for preparation of the sample as given in  parts 4.1.1 thru 4.1.4 of the
           Atomic Absorption Methods section of this manual have been found to be satisfactory.

Instrumental Parameters (General)
     1.    Iron hollow cathode lamp
     2.    Wavelength: 248.3 nm
     3.    Fuel: Acetylene
     4.    Oxidant: Air
     5.    Type of flame: Oxidizing

Analysis Procedure
     1.    For analysis procedure and calculation, see "Direct Aspiration", part 9.1 of the Atomic
           Absorption Methods section of this manual.
 Approved for NPDES
 Issued 1971
 Editorial revision  1974  and 1978
                                        G-12

-------
Notes
     1.    The following lines may also be used:
           248.8 nm Relative Sensitivity 2
           271.9 nm Relative Sensitivity 4
           302.1 nm Relative Sensitivity 5
           252.1 nm Relative Sensitivity 6
           372.0 nm Relative Sensitivity 10
     2.    Data to be reported into STORET must be reported as ug/1.
     3.    The 1,10-phenanthroline colorimetric method may also be used (Standard Methods,
           14th Edition, p. 208).
     4.    For concentrations of iron below 0.05  mg/1, either the Special Extraction Procedure
           given in part 9.2 of the Atomic Absorption Methods section or the furnace procedure,
           Method 236.2, is recommended.

Precision and Accuracy
     1.    An interlaboratory study on trace metal analyses by atomic absorption was conducted by
           the Quality Assurance and Laboratory Evaluation Branch of EMSL.  Six synthetic
           concentrates containing varying levels of aluminum, cadmium, chromium, copper, iron,
           manganese, lead and zinc were added to natural water samples. The statistical results for
           iron were as follows:
                                                          Standard
 Number        True Values          Mean  Value            Deviation            Accuracy as
 Of Labs          ug/liter              ug/liter              tig/liter              % Bias

  82              840                 855                 173                    1.8
  85              700                 680                 178                   -2.8
  78              350                 348                 131                   -0.5
  79              438                 435                 183                   -0.7
  57               24                  58                  69                  141
  54               10                  48                  69                  382
                                               G-13

-------
                                        LEAD
              Method  239.1  (Atomic Absorption,  direct aspiration)

                                                        STORET  NO.  Total 01051
                                                                     Dissolved 01049
                                                                    Suspended 01050

Optimum Concentration Range:   1-20 mg/1 using a wavelength of 283.3 nm
Sensitivity:     0.5 mg/1
Detection Limit:      0.1 mg/1

Preparation of Standard Solution
     1.   Stock Solution: Carefully weigh 1.S99 g of. lead nitrate, Pb(NO3)2 (analytical reagent
          grade), and dissolve in deionized distilled water. When solution is complete acidify with
          10 ml redistilled HNO3 and dilute to 1 liter with deionized distilled water. 1 ml =  1 mg
          Pb( 1000 mg/1).
     2.   Prepare dilutions of the stock solution to be used as calibration standards at the time of
          analysis. The calibration standards should be prepared using the same type of acid and at
          the same concentration as will result in the sample to be analyzed either directly or after
          processing.

Sample Preservation
     1.   For sample handling and preservation, see  part 4.1 of the Atomic Absorption Methods
          section of this manual.

Sample Preparation
     1.   The procedures for preparation of the sample as given in parts  4.1.1 thru 4.1.4 of the
          Atomic Absorption Methods section of this manual have been found to be satisfactory.

Instrumental Parameters (General)
      1.   Lead hollow cathode lamp
     2.   Wavelength: 283.3 nm
     3.   Fuel: Acetylene
     4.   Oxidant: Air
     5.   Type of flame: Oxidizing

Analysis Procedure
      1.    For analysis procedure and calculation, see "Direct Aspiration", part 9.1 of the Atomic
           Absorption  Methods section of this manual.
 Approved for NPDES and SDWA
 Issued  1971
 Editorial revision  1974  and 1978
                                        G-14'

-------
Notes
     1.    The analysis of this metal is exceptionally sensitive to turbulence and absorption bands in
           the flame. Therefore, some care should be taken to position the light beam in the most
           stable, center portion of the flame. To do this, first adjust the burner to maximize the
           absorbance reading with a lead standard. Then, aspirate a water blank and make minute
           adjustments in the burner alignment to minimize the signal.
     2.    For levels of lead below 200 i/g/1, either the Special Extraction Procedure given in part
           9.2 of the Atomic Absorption Methods section or the furnace technique, Method 239.2,
           is recommended.
     3.    The following lines may also be used:
           217.0 nm Relative Sensitivity 0.4
           261.4 nm Relative Sensitivity 10
     4.    For quality control  requirements and optional recommendations for use in drinking
           water analyses, see part 10 of the Atomic Absorption Methods section of this manual.
     5.    Data to be entered into STORET must be reported as ug/1.
Precision and Accuracy
      1.    An interlaboratory study on trace metal analyses by atomic absorption was conducted by
           the  Quality Assurance and Laboratory  Evaluation  Branch of EMSL.  Six synthetic
           concentrates containing varying levels of aluminum, cadmium, chromium, copper, iron,
           manganese, lead and zinc were added to natural water samples. The statistical results for
           lead were as follows:
                                                          Standard
 Number        True Values          Mean Value           Deviation            Accuracy as
 of Labs           iig/liter              ug/liter              iig/liter              % Bias

   74              367                  377                  128                  2.9
   74              334                  340                  111                  1.8
   64              101                  101                   46                 -0.2
   64               84                  85                   40                  1.1
   61               37                  41                   25                  9.6
   60               25                  31                   22                 25.7
                                              G-15

-------
                                   MANGANESE
              Method  243.1 (Atomic Absorption, direct aspiration)

                                                        STORET NO.  Total 01055
                                                                     Dissolved 01056
                                                                   Suspended 01054

Optimum Concentration Range:   0.1-3 mg/1 using a wavelength of 279.5 nm
Sensitivity:     0.05 mg/1
Detection Limit:      0.01 mg/1

Preparation of Standard Solution
     1.    Stock Solution: Carefully weigh 1.000 g of manganese metal (analytical reagent grade)
          and dissolve in 10 ml of redistilled HNO3. When solution is complete, dilute to  1 liter
          with 1%(V/V)HC1.1ml = 1 mgMn (1000 mg/1).
     2.    Prepare dilutions of the stock solution to be used as calibration standards at the time of
          analysis. The calibration standards should be prepared using the same type of acid and at
          the same concentration as will result in the sample to be analyzed either directly or after
          processing.

Sample Preservation
     1.    For sample handling and preservation, see part 4.1 of the Atomic Absorption Methods
          section of this manual.

Sample Preparation
     1.    The procedures for preparation of the sample as given in parts 4.1.1 thru 4.1.4  of the
          Atomic Absorption Methods section of this manual have been found to be satisfactory.

Instrumental Parameters (General)
     1.    Manganese hollow cathode lamp
     2.    Wavelength: 279.5 nm
     3.    Fuel: Acetylene
     4.    Oxidant: Air
     5.    Type of flame: Oxidizing

Analysis Procedure
     1.    For analysis procedure and calculation, see "Direct Aspiration", part 9.1 of the Atomic
          Absorption Methods section of this manual.
Approved for NPDES
Issued 1971
Editorial revision 1974 and 1978
                                    G-16

-------
Notes
     1.    For levels of manganese below 25 ug/1, either the furnace procedure. Method 243.2, or
           the Special Extraction Procedure given in part 9.2 of the Atomic Absorption Methods
           section is recommended. The extraction is carried out at a pH of 4.5 to 5. The manganese
           chelate is very unstable and the analysis must be made without delay to prevent its re-
           solution in the aqueous phase.
     2.    The following line may also be used:
           403.1 nm Relative Sensitivity 10.
     3.    Data to be entered into STORET must be reported as ug/1.
     4.    The persulfate colorimetric method may also be used (Standard Methods, 14th Edition,
           p225).

Precision and Accuracy
     1.    An interlaboratory study on trace metal analyses by atomic absorption was conducted by
           the  Quality Assurance  and Laboratory Evaluation Branch  of EMSL. Six synthetic
           concentrates containing varying levels of aluminum, cadmium, chromium, copper, iron,
           manganese, lead and zinc were added to natural water samples. The statistical results for
           manganese were as follows:
                                                         Standard
Number        True Values         Mean Value           Deviation           Accuracy as
of Labs          ug/liter              ug/liter              ug/liter              % Bias

  77             426                  432                  70                    1.5
  78             469                  474                  97                    1.2
  71              84                   86                  26                    2.1
  70             106                  104                  31                   -2.1
  55              II                   21                  27                   93
  55              17                   21                  20                   22
                                               G-17

-------
                                       NICKEL
               Method 249.1 (Atomic  Absorption, direct aspiration)

                                                         STORE! NO. Total 01067
                                                                      Dissolved 01065
                                                                     Suspended 01066

Optimum Concentration Range:   0.3-5 mg/1 using a wavelength of 232.0 nm
Sensitivity:      0.15 mg/1
Detection Limit:      0.04 mg/1

Preparation of Standard Solution
      1.    Stock Solution: Dissolve 4.953 g  of nickel nitrate, Ni(NO3)2»6H2O (analytical reagent
           grade) in deionized distilled water. Add 10 ml of cone, nitric acid and dilute to 1  liter
           with deionized distilled water. 1 ml = 1 mg Ni (1000 mg/1).
      2.    Prepare dilutions of the stock nickel solution to be used as calibration standards at the
           time of analysis. The calibration standards should be prepared using the same type of
           acid and at the same concentration as will result in the sample to be analyzed either
           directly or after processing.

Sample Preservation
      1.    For  sample handling and preservation, see part 4.1 of the Atomic Absorption Methods
           section of this manual.

Sample Preparation
      1.    The  procedures for preparation of the sample as given in parts 4.1.1 thru 4.1.4 of the
           Atomic Absorption Methods section of this manual have been found to be satisfactory.

Instrumental Parameters (General)
      1.    Nickel hollow cathode lamp
     2.    Wavelength: 232.0 nm
     3,    Fuel: Acetylene
     4.    Oxidant: Air
     5.    Type of Flame: Oxidizing

Analysis Procedure
      1.    For analysis procedure and calculation, see "Direct Aspiration", part 9.1 of the Atomic
           Absorption Methods section of this manual.
Approved for  NPDES
Issued  1974
Editorial revision 1978
                                     G-18

-------
Interferences
      1.    The 352.4 nm wavelength is less susceptible to spectral interference and may be used.
           The calibration curve is more linear at this wavelength; however, there is some loss of
           sensitivity.

Notes
      1.    For levels of nickel below 100 ug/1, either the Special Extraction Procedure, given in
           part 9.2 of the Atomic Absorption Methods section or the furnace technique, Method
           249.2, is recommended.
      2.    Data to be entered into STORET must be reported as ug/1.
      3.    The heptoxime method may also be used (Standard Methods, 14th Edition, p 232).

Precision and Accuracy
      1.    In a single laboratory (EMSL), using  a mixed industrial-domestic waste effluent at
           concentrations of 0.20,1.0 and 5.0 mgNi/1, the standard deviations were ±0.011,  ±0.02
           and  ±0.04, respectively. Recoveries at  these  levels were 100%, 97%  and  93%,
           respectively.
                                                G-19

-------
                                       SILVER

              Method 272.1  (Atomic Absorption, direct aspiration)

                                                         STORE! NO. Total  01077
                                                                      Dissolved  01075
                                                                     Suspended  01076

Optimum Concentration Range:   0.1-4 mg/1 using a wavelength of 328.1 nm
Sensitivity:      0.06 mg/1
Detection Limit:       0.01 mg/1

Preparation of Standard Solution
      1.    Stock Solution: Dissolve 1.575 g  of AgttO3 (analytical reagent grade) in deionized
           distilled water, add 10 ml cone. HNO3 and make up to 1 liter. 1 ml =  1 mg Ag (1000
           mg/1).
      2.    Prepare dilutions of the stock solution to be used as calibration standards at the time of
           analysis. The calibration standards should be prepared using nitric acid and at the same
           concentration as  will result in the sample to be  analyzed either directly  or after
           processing.
      3.    Iodine Solution, 1 N: Dissolve 20 grams of potassium iodide, KI (analytical reagent
           grade) in  50 ml of deionized distilled water, add 12.7 grams of iodine, I2  (analytical
           reagent grade) and dilute to 100 ml. Store in a brown bottle.
      4.    Cyanogen Iodide (CNI) Solution: To 50  ml of deionized distilled water add 4.0 ml cone.
           NH4OH, 6.5 grams KCN, and 5.0 ml of  1.0 N I2 solution. Mix and dilute to 100 ml with
           deionized distilled water. Fresh solution should be prepared every two weeks."'

Sample Preservation
      1.    For sample handling and preservation, see part 4.1 of the Atomic Absorption Methods
           section of this manual.

Sample Preparation
      1.    The procedures for preparation of the sample as given in parts 4.1.1 thru 4.1.3 of the
           Atomic Absorption Methods section of  this manual  have been found to be satisfactory;
           however,  the residue must be taken up in dilute nitric  acid rather than  hydrochloric to
           prevent precipitation of AgCl.
 Approved  for NPDES and SOW A
 Issued  1971
 Editorial revision 1974
 Technical  revision  1978
                                          G-20

-------
Instrumental Parameters (General)
     1.    Silver hollow cathode lamp
     2.    Wavelength: 328.1 nm
     3.    Fuel: Acetylene
     4.    Oxidant: Air
     5.    Type of flame: Oxidizing

Analysis Procedure
     1.    For the analysis procedure and the calculation, see "Direct Aspiration", part 9.1 of the
          Atomic Absorption Methods section of this manual.

Motet
     I.    For levels of silver below 30 tig/1, either the Special Extraction Procedure, given in part
          9.2 of the Atomic Absorption Methods section or the furnace procedure, Method 272.2,
          is recommended.
     2.    Silver nitrate standards are light sensitive. Dilutions of the stock should be discarded
          after use as concentrations below 10 mg/1 are not stable over long periods of time.
     3.    If absorption to container walls or the formation of AgCl is suspected, make the sample
          basic using cone. NH4OH and add 1 ml of (CNI) solution per 100 ml of sample. Mix the
          sample and allow to stand for 1 hour before proceeding with the analysis.1"
     4.    The 338.2 nm wavelength may also be used. This has a relative sensitivity of 2.
     5.    Data to be entered into STORET must be reported as ug/1.
Precision and Accuracy
      1.    In a round-robin study reported by Standard Methods, a synthetic sample containing SO
           ug Ag/1 was analyzed by SO laboratories with a reported standard deviation of ±8.8 and
           a relative error 10.6%.

References •
      1.    "The Use of Cyanogen Iodide (CNI) as a Stabilizing Agent for Silver in Photographic
           Processing Effluent Sample", Owerbach, Daniel, Photographic Technology Division,
           Eastman Kodak Company, Rochester, N.Y. 146SO.
      2.    Standard Methods for Examination of Water and Wastewater, 14th Edition, p. 148,
           Method 301A.
                                                G-21

-------
                                        ZINC

              Method  289.1 (Atomic Absorption, direct  aspiration)

                                                        STORE!  NO. Total 01092
                                                                     Dissolved 01090
                                                                    Suspended 01091

Optimum Concentration Range:  0.05-1 mg/1 using a wavelength of 213.9 nm
Sensitivity:     0.02 mg/1
Detection Limit:       0.005 mg/1

Preparation of Standard Solution
      1.    Stock Solution: Carefully weigh 1.00 g of zinc metal (analytical reagent grade) and
           dissolve cautiously in 10 ml HNO3. When solution is complete make up to 1 liter with
           deionized distilled water. 1 ml = 1 mg Zn (1000 mg/1).
      2.    Prepare dilutions of the stock solution to be used as calibration standards at the time of
           analysis. The calibration standards should be prepared using the same type of acid and at
           the same concentration as will result in the sample to be analyzed either directly or after
           processing.

Sample Preservation
      1.    For sample handling and preservation, see part 4.1 of the Atomic Absorption Methods
           section of this manual.

Sample Preparation
      1.    The procedures for preparation of the sample as given in parts 4.1.1 thru 4.1.4 of the
           Atomic Absorption Methods section of this manual have been found to be satisfactory.

Instrumental Parameters
      1.    Zinc hollow cathode lamp
      2.    Wavelength: 213.9 nm
      3.    Fuel: Acetylene
      4.    Oxidant: Air
      5.    Type of flame: Oxidizing

 Analysis Procedure
      1.    For the analysis procedure and the calculation, see "direct aspiration" part 9.1 of the
           Atomic Absorption Methods section of this manual.
 Approved for  NPDES
 Issued 1971
 Editorial revision 1974
                                        G-22

-------
Notes
     1.     High levels of silicon may interfere.
     2.     The air-acetylene flame absorbs about 25% of the energy at the 213.9 nm line.
     3.     The sensitivity may be increased by the use of low-temperature flames.
     4.     Some sample container cap liners can be a source of zinc contamination. To circumvent
           or avoid this problem, the use of polypropylene caps is recommended.
     5.     The dithizone colorimetric method may also be used (Standard Methods, 14th Edition, p
           26S).
     6.     For concentrations of zinc below 0.01  mg/1, either the Special Extraction  Procedure
           given in part 9.2 of the Atomic Absorption Methods section or the furnace procedure,
           Method 289.2, is recommended.
     7.    Data to entered into Storet must be reported as ug/1.

Precision and Accuracy
      1.    An interlaboratory study on trace metal analyses by atomic absorption was conducted by
           the Quality Assurance and Laboratory Evaluation Branch of EMSL. Six synthetic
           concentrates containing varying levels of aluminum, cadmium, chromium, copper, iron,
           manganese, lead and zinc were added to natural water samples. The statistical results for
           zinc were as follows:
                                                          Standard
 Number        True Values          Mean  Value           Deviation           Accuracy as
 of Labs          ug/liter              ug/liter              tig/liter              % Bias

   86              281                 284                  97                   1.2
   89              310                 308                 114                  -07
   82               56                  62                  28                  11.3
   81               70                  75                  28                   6.6
   62                7                  22                  26                 206
   61               11                  17                  18                  56.6
                                               G-23

-------
BENTH1C JNFAUHA

-------
SEDIMENT BiOASSAYS

-------
FINAL REPORT
TC-3991-04                       PugetSound Estuary Program
RECOMMENDED PROTOCOLS FOR
CONDUCTING LABORATORY BIOASSAYS
ON PUGET SOUND SEDIMENTS
Prepared by:
TETRA TECH, INC.
and
E.V.S. CONSULTANTS, INC.
Prepared for:
U.S. ENVIRONMENTAL PROTECTION AGENCY
Region 10 - Office of Puget Sound
Seattle, WA
May, 1986
TETRA TECH, INC.
11820 Northup Way
Bellevue, WA 98005

-------
                                 CONTENTS
LIST OF TABLES                                                            v



ACKNOWLEDGEMENTS                                                          v



INTRODUCTION                                                              l



BIOASSAY SELECTION                                                        4



FIELD COLLECTION OF SUBTIDAL SURFICIAL SEDIMENTS                         11



     DESIGN OF SAMPLER                                                   U



     PENETRATION DEPTH                                                   12



     OPERATION OF SAMPLER                                                13



     SAMPLE ACCEPTABILITY CRITERIA                                       13



     SAMPLE COLLECTION                                                   14



     SAMPLE HOMOGENIZATION                                               15



     CONCURRENT COLLECTION OF SEDIMENT CHEMISTRY AND BIOASSAY SAMPLES    15



GENERAL QUALITY ASSURANCE/QUALITY CONTROL GUIDELINES                     16



     NEGATIVE CONTROLS                                                   16



     POSITIVE CONTROLS                                                   16



     TEST ORGANISMS                                                      16



     REFERENCE TEST SAMPLES                                              16



     BLIND TESTING                                                       17



     MAINTENANCE/MEASUREMENT OF WATER QUALITY                            17



     STANDARD LABORATORY PROCEDURES                                      17



AMPHIPOD SEDIMENT BIOASSAY                                               18



     USE AND LIMITATIONS                                                 18
                                   11

-------
     FIELD PROCEDURES                                                    19

          Collection                                                     19
          Processing                                                     19

     LABORATORY PROCEDURES                                               19

          Test Animals                                                   20
          Control  Sediment                                               21
          Test Sediment                                                  21
          Bioassay Seawater                                              22
          Facilities and Equipment                                       22
          Bioassay Procedure                                             23
          Experimental Design                                            25

     DATA REPORTING REQUIREMENTS                                         26

BIVALVE LARVAE SEDIMENT BIOASSAY                                         27

     USE AND LIMITATIONS                                                 27

     FIELD PROCEDURES                                                    28

          Collection                                                     28
          Processing                                                     28

     LABORATORY PROCEDURES                                               28

          Bioassay Species                                               29
          Bioassay Sediment                                              29
          Bioassay Seawater                                              29
          Facilities and Equipment                                       30
          Bioassay Procedure                                             30
          Controls                                                       32

     DATA REPORTING REQUIREMENTS                                         32

ANAPHASE ABERRATION SEDIMENT BIOASSAY                                    33

     USE AND LIMITATIONS                                                 33

     FIELD PROCEDURES                                                    33

          Collection                                                     33
          Processing                                                     33

     LABORATORY PROCEDURES                                               34

          Cell Cultures                                                  34
          Sediment Extraction                                            34
          Culture Conditions                                             36
          Bioassay Procedure                                             37
          Controls                                                       38


                                  111

-------
     DATA REPORTING REQUIREMENTS                                         38

MICROTOX SEDIMENT BIOASSAY (ORGANIC EXTRACT)                             40

     USE AND LIMITATIONS                                                 40

     FIELD PROCEDURES                                                    41

          Collection                                                     41
          Processing                                                     41

     LABORATORY PROCEDURES                                               41

          Facilities and Equipment                                       41
          Sediment Extraction                                            41
          Bioassay Procedure                                             43
          Controls                                                       44

     DATA REPORTING REQUIREMENTS                                         44

MICROTOX SEDIMENT BIOASSAY (SALINE EXTRACT)                              45

     USE AND LIMITATIONS                                                 45

     FIELD PROCEDURES                                                    45

          Collection                                                     45
          Processing                                                     45

     LABORATORY PROCEDURES                                               46

          Preparation of Sediment Extract                                46
          Bioassay Procedure                                             46
          Controls                                                       47

     DATA REPORTING REQUIREMENTS                                         48

OTHER PROMISING TECHNIQUES                                               49

REFERENCES                                                               51
                                    1v

-------
                                  TABLES


Number                                                                Page

  1       Contributors to the sediment bioassay protocols                 2

  2       Summary  of  laboratory bioassay tests with marine  sediments      5

  3       Applicability and relevance of sensitive laboratory
          sediment bioassay tests used in more than one  study             8

  4       Categorization of selected sediment bioassays  by  type of
          test and type of effect                                       10
                             ACKNOWLEDGEMENTS
     This chapter was prepared by Tetra  Tech, Inc., under  the direction
of Dr. Scott Becker,  for  the U.S. Environmental Protection  Agency  in partial
fulfillment of Contract No.  68-03-1977.   Dr. Thomas  Sinn of Tetra Tech
was the Program Manager.  Mr.  John  Underwood  and  Dr.  John Armstrong of
U.S.  EPA  were the Project Officers.  The principal  author  of  this chapter
was Dr. Peter Chapman of  E.V.S.  Consultants, Inc.

-------
                                              Laboratory Sediment Bioassays
                                                              Introduction
                                                                  May 1986
                               INTRODUCTION


     This document  presents recommended methods for  conducting the following
laboratory sediment bioassays in Puget Sound:

     •    Amphipod  bioassay

     •    Bivalve larvae bioassay

     •    Anaphase  aberration bioassay

     •    Microtox  bioassay

               Organic extraction
               Saline extraction.

Each  recommended  method  is  based on the results of a  workshop and written
reviews by representatives from most  organizations that fund or  conduct
environmental  studies in Puget Sound (Table 1).   The purpose of developing
these recommended protocols is  to encourage all Puget Sound investigators
conducting monitoring programs, baseline surveys,  and intensive investigations
to use standardized methods whenever possible.  If  this goal is achieved,
most  data collected in  the  Sound will  be directly comparable and thereby
capable of being  integrated into a  sound-wide  database.   Such a database
is necessary for  developing and maintaining a comprehensive water quality
management program  for Puget  Sound.

     The  sediment  bioassays considered in this document  are the tests used
most frequently by  a variety  of Puget Sound investigators.   However,  other
bioassays have been used or are currently being  developed  for use  in Puget
Sound.  Several of  these  additional tests are described  in the final section
of this document  (see Other Promising Techniques).

     Each  recommended  protocol  describes  the use  and  limitations of the
respective variable; the  field  collection and  processing methods;  and  the
laboratory analytical,  quality assurance/quality control (QA/QC),  and data
reporting procedures.  In  developing each protocol,  it was recognized  that
the  field of sediment bioassays is relatively new and is  expanding  rapidly.
The loose-leaf format of  this document will allow modification of the recom-
mended protocols,  if necessary, and inclusion of additional  protocols.

     Although  the  following protocols are  recommended for  most studies
conducted in Puget  Sound,  departures  from these  methods may be necessary
to meet the special requirements of individual projects.  If such departures
are made, however,  the funding  agency or investigator should be  aware  that
the  resulting data may not be  comparable with most other  data of that kind.

                                   1

-------
         TABLE 1.  CONTRIBUTORS TO THE SEDIMENT BIOASSAY PROTOCOLS
John Armstrong3
Richard Batiuk
Scott Becker3
Alice Benedict3
Jim Buckley3
Rick Cardwell3
Gary Chapman   .
Peter Chapman3'0
Joe Cummins3
Paul Dinnel
Bruce Duncan3
Jim Fisher3
Burt Hamner3
Dave Jamison3
Dave Kendall3
Dick Kocan3
Mary Lou Mills3
Phillip Oshida
Diane Robbins
Pat Romberg3
Brian Ross
Norman Rubinstein
Mike Schiewe3
Margaret Stinson3
Jerry Stober3
John Strand3
Rick Swartz
Dave Terpening3
Jim Thornton3
Milton Tunzi
Don Weitkamp3
Ron Westley3
Les Williams3
Jack Word3
U.S.  EPA
U.S.  EPA
Tetra Tech
NOAA
Metro
Envirosphere Co.
U.S.  EPA
EVS Consultants
U.S.  EPA
Univ. of Washington
U.S.  EPA
Weyerhaeuser Co.
U.S.  COE
Wash. Dept. of Natural Resources
     COE
      of Washington
      Dept. of Fisheries
U.S.
Univ
Wash
U.S.  EPA
Invert Aid
Metro
U.S.  EPA
U.S.  EPA
NOAA
Wash. Dept. of Ecology
Univ. of Washington
Battelle Northwest
U.S.  EPA
U.S.  EPA
Wash. Dept. of Ecology
U.S.  EPA
Parametrix
Wash. Dept. of Fisheries
Tetra Tech
Evans-Hamilton
  Attended the workshop held on October 22, 1985.
  Workshop moderator.

-------
                                              Laboratory Sediment Bioassays
                                                              Introduction
                                                                  May 1986

In some  instances, data  collected using different methods may be compared
if the methods  are  intercalibrated adequately.

     Before protocols for  specific bioassays  are described, sections are
presented on 1)  the criteria used to identify the tests  considered  in  this
document,  2)  protocols  for field  collection of surficial test sediments,
and 3) general  QA/QC procedures that apply to all  sediment bioassays.

-------
                                             Laboratory Sediment  Bioassays
                                                              Introduction
                                                                  May  1986
                            BIOASSAY SELECTION


     A large number  of sediment  bioassays have  been  developed and  used
in recent  years  (Table 2).  Many of these  tests have been  used  to evaluate
the toxicity of  Puget Sound sediments.  Eight tests (Table 3) were selected
for further  consideration based on the  following criteria:

     •    Sensitivity - each  test  has detected  biological effects
          in a variety of sediments

     •    Usage - each  test has been  used  in  more than one study  in
          Puget  Sound.

     Of the eight  sediment bioassays  selected for consideration, four  were
identified as suitable for general application in Puget  Sound  for reasons
outlined  in Table 3.  These four tests are  categorized  in Table 4 in terms
of kind of test  and kind of effect measured.  The other  tests  identified
in Table  3, together with some of the  additonal  tests mentioned  in Table 2,
are considered as other promising techniques  at the end of this chapter.

-------
              TABLE 2.  SUMMARY OF LABORATORY  BIOASSAY TESTS
                          WITH MARINE SEDIMENTS
        Taxon
    Investigators
ACUTE LETHAL BIOASSAYS
  Phytoplankton:
          two species test
  Annelid worms:
     *    Monopylephorus  cuticulatus
     *    Glycinde picta
  Crustaceans:
     (Copepods)
     *    Acartia tonsa
     *    Tigriopus sp.
     (Amphipods)
     *    Eogammarus confervicolus
     *    Eohaustorius washingtonianus
     *    Grandifoxus  grandis
    T*    Rhepoxynius  abronius
     (Cumaceans)
          several  species
     (Shrimp)
     *    Palaemonetes pugio
Wurster (1982)

Chapman et al.  (198Za)
Swartz et al.  (1979)
Shuba et al.  (1978)
Shuba et al.  (1978)

Chapman et al.  (1982a)
Ott et al. (in  prep.)
Pierson et al.  (1983)
Swartz  et  al  .  (1979;  1981;
1982; 1985a,b); Ott  et al .  (in
prep.); Chapman et al.  (1982a,b;
1984); Chapman  and Fink (1983);
Chapman  and   Barlow  (1984);
Malins  et al.   (1985);  Tetra
Tech (1985);  Battelle  (1986)
Swartz et al.  (1979)
Shuba et al. (1978)

-------
TABLE 2.   (Continued)
  Bivalve Molluscs:

     *    Macoma  inquinata

     *    Protothaca  staminea

     *    Rangia  cuneata

          Mya arenaria
  Fish:

     *
          Gasterosteus  aculeatus
          (threespine  stickleback)

          Oncorhynchus  kisutch
          (coho salmon)

          Oncorhynchus  tshawytscha
          (Chinook  salmon)

          Fundulus  heteroclitus
          (mummichog)

          Leiostomus  xanthurus
          (spot)

SUBLETHAL BIOASSAYS
          Respiration rate of the
          oligochaete worm,
          Monopylephorus  cuticulatus

          Oyster larvae.  Crassostrea gigas,
          partial  life-cycle bioassay
          Surf smelt, Hypomesys  pretiosus
          pretiosus, partial  life-cycle
          Bioassay

          Full life-cycle bioassay with
          the polychaete worm,
          Capitella capitata
Swartz et al.  (1979)

Swartz et al.  (1979)

Shuba et al.  (1978)

Tsai et al.  (1979)



Chapman et al.  (1982a)


Legore and DesVoigne  (1973)


Pierson et al.  (1983)
Tsai et al.  (1979);
Hargis et al.  (1984)

Tsai et al.  (1979);
Hargis et al.  (1984)
Chapman  (in press);  Chapman
et  al.  (1982a,b; 1984;  1985);
Chapman and Fink (1983)

Schink  et al .  (1974); Cummins
et  al .  (1976);  Chapman  and
Morgan  (1983);  Chapman  et
al .  (1983;  1985); Pierson
et  al.  (1983); Mai ins  et  al .
(1985);  Tetra  Tech (1985);
Battelle  (1986)

Chapman  et al.  (1983);  Malins
et  al. (1985)
 Chapman et al.  (1983);
 Chapman and Fink (1984)

-------
TABLE 2.   (Continued)
    T*
Histopathological  measures of
toxicity to English sole,
Parophrys vetulus, with 3 mo.
sediment exposure

Phytotoxicity to marsh plant
Cyperus esculentus

Sea urchin fertilization assay
with Strongylocentrotus
purpuratus

Full life-cycle bioassays with
nematodes Chromadorina germanica
and Diploeaimella  punicea

Partial life-cycle bioassays
with the copepod Tigriopus
californicus
CYTOTOXIC/GENOTOXIC BIOASSAYS

    T*    Mitotic abnormalities (anaphase
          aberrations) and cell prolifera-
          tion in vitro

MICROBIAL ACTIVITY BIOASSAYS

    T*    Microtox, Photobacterium
          phosphoreum, bacterial
          luminescence
                                             McCain  et al.  (1982)
                                             Lee  et  al.  (1982)
                                             Ross  et  al.  (1984)
                                             Malins et  al.  (1985)
                                             Tietjen  and  Lee  (1984)
Misitano  (1983) ;  Mai ins  et
al. (1985)
                                   Chapman  et  al.  (1982a,b;  1983;
                                   1984;  1985);  Landolt  and  Kocan
                                   (1984a,b)
                                   Williams et al.  (in  press);
                                   Schiewe  et al.  (1985);  Malins
                                   et al.  (1985)
* Indicates used in Puget Sound.

T Sensitive method used in more than one study.

-------
           TABLE 3.  APPLICABILITY AND RELEVANCE  OF  SENSITIVE LABORATORY
                SEDIMENT BIOASSAY TESTS USED IN MORE THAN ONE STUDY
Bioassay
                Type'
Acceptable  for
Immediate Use
in Puget Sound?
         Comment
Amphipod  10-day test
(Rhepoxynius abronius)
                 L.S
     Yes
Oligochaete  respiration
rate measurements
(Monopylephorus cuticulatus)
                               No
Bivalve larvae 48-h test
(Crassostrea  gigas,
Mytilus edulis
                 L,S
     Yes
gigas.
Echinoderm sperm test
                 L.S
     No
Surf smelt partial life-
cycle test (Hypomesus
pretiosus  pretiosus)
                L.S.C
     No
Copepod partial  life-cycle    L,S,C
test (Tigriopus  californicus)
                               No
Simple test, reproducible,
of ecological  relevance,
applicability  verified
in  numerous  studies,
extensive  usage  to date
in Puget  Sound

Provides  useful  information,
but requires  specialized
equipment and expertise
not generally  available
(Chapman  et al. 1985)

Simple test,  indicator
of toxicity,  relevance
and applicability verified
in several independent
studies, extensive usage
to date in  Puget Sound

Simple test, can be conducted
year-round by using different
echinoderm species,  but
interfering factors cannot
yet  be eliminated (Ross
et al.  1984;  Malins  et
al.   1985; Dinnel  and
Stober 1985)

Smelt  eggs not available
year-round,  results confounded
by  excessive natural
var iab i1ity  (Chapman
et al. 1983, 1985)

Simple test, can be conducted
year-round, but procedure
has not been fully validated
(D. Misitano,  NMFS, pers.
ccmm.; P. Chapman,  E.V.S.
Consultants, unpub. data)
                                          8

-------
TABLE  3.   (Continued)
In vitro  anaphase aberration
tests with rainbow trout
gonad cells  (Salmo gairdneri)
Yes
Microtox,  bacterial
luminescence bioassay
(Photobacterium phosphoreum)
Yes
One  of  the  few methods
to measure possible genotoxic/
cytotoxic effects, moderate
usage to  date  in Puget
Sound, requires  a higher
level of expertise than
other  tests  but can  be
conducted  by more than
one  laboratory  in  Puget
Sound

One  of  the  few methods
to measure possible effects
on bacteria, shows good
promise, moderate usage
to date  in Puget Sound
a L = Lethal; S = Sublethal;  C  = Chronic; G = Genotoxic.

-------
          TABLE 4.   CATEGORIZATION  OF  SELECTED SEDIMENT BIOASSAYS
                    BY KIND OF TEST AND KIND OF  EFFECT


                                       Kind  of Effect3
Kind of Test                  Lethal      Sublethal     Genotoxic

Solid and liquid phase:

  Rhepoxynius abronius           X            X

  Bivalve larvae                 X            X

Extracts:

  Microtox                                   X

  Anaphase aberrations                                      X


a None  of these tests  measures chronic effects (i.e., defined as longer
term than 10 days,  involving partial  life-cycle  testing).
                                   10

-------
                                             Laboratory Sediment Bioassays
                                                       Sediment Collection
                                                                 May 1936
             FIELD COLLECTION OF SUBTIDAL SURFICIAL  SEDIMENTS


     This section describes the protocols required to collect  an  acceptable
subtidal  surficial sediment sample  for  subsequent analysis by  a  laboratory
sediment  bioassay.   This  subject has generally been neglected  in the past
and sampling  crews have  been  given relatively  wide latitude  in deciding
how to collect samples.   However, because sample collection  procedures
influence the results of all subsequent  laboratory  and data  analyses,  it
is critical  that samples  be  collected using  acceptable and standardized
techniques.

DESIGN OF SAMPLER

     In Puget  Sound, the most common  sampling device for subtidal  surficial
sediments is the modified van Veen bottom  grab.  However,  a  variety of
coring devices (e.g., box corer,  Kasten corer) are also used.  The  primary
criterion for  an adequate sampler is that  it consistently collect  undisturbed
samples to the required depth below the sediment surface without contaminating
the samples.   An  additional criterion  is that  the sampler  can  be handled
properly on  board the  survey vessel.  An otherwise acceptable  sampler may
yield inadequate  sediment  samples  if  it is  too large, heavy,  or awkward
to be handled properly.

     Collection of undisturbed sediment requires that the sampler:

     •    Create  a minimal  bow wake when  descending

     •    Form a  leakproof  seal when the  sediment  sample is taken

     •    Prevent winnowing  and  excessive sample disturbance  when
          ascending

     •    Allow easy  access to the sample surface.

Most  modified van Veen  grabs have open upper faces that are fitted with
rubber flaps.  Upon  descent, the  flaps are forced open  to  minimize the
bow wake, whereas  upon ascent, the flaps  are  forced  closed to prevent sample
winnowing.  Some box  corers have  solid  flaps  that are  clipped open upon
descent, and snap  shut after the corer is triggered.  Although most samplers
seal adequately when  purchased, the wear  and  tear of repeated field use
eventually reduces  this  sealing  ability.  A  sampler should therefore oe
monitored constantly  for sample leakage.   If  unacceptable  leakage occurs,
the  sampler should be repaired or replaced.  If a sampler is to be borrowed
or leased for  a project,  its sealing ability should be  confirmed prior


                                   11

-------
                                             Laboratory  Sediment Bioassays
                                                      Sediment Collection
                                                                 May 1986


to sampling.   Also, it is prudent  to have a backup  sampler on board the
survey vessel  if the primary sampler  begins leaking during  a cruise.

     The required penetration  depth below the sediment surface is a function
of the desired sample depth  (see Penetration  Depth).  Generally,  it  is
better  to  penetrate below the  desired sample  depth to minimize  sample
disturbance when the sampling device closes.   Penetration depth of  most
sampling devices  varies with  sediment character;  it is greatest in fine
sediments and  least  in coarse  sediments.  Sampling devices generally  rely
upon  either  gravity or a  piston mechanism to penetrate  the sediment.  In
both cases,  penetration depth  can  be modified  by adding or subtracting
weight  from  the  samplers.   Thus,  it is optimal to use  a sampler that has
a means of weight adjustment.   If a  sampler cannot consistently achieve
the desired penetration depth, an alternate device should  be used.

     Once the  sampler is secured  on board the survey vessel, it is essential
that the surface of the sample be made accessible without disturbing the
sample.  Generally, samplers have  hinged flaps  on their  upper face for
this purpose.   The openings  in  the  upper face  of the sampler  should  be
large  enough  to  allow easy  subsampling of the  sediment surface.   If an
opening is too small, the  sample  may be  disturbed as the field member struggles
to take a subsample.

PENETRATION DEPTH

     For characterizing the toxicity of surficial sediments in Puget Sound,
it is  recommended that the upper  2 cm of  sediment be evaluated.  When collecting
the upper 2  cm of sediment,  it is  recommended that a minimum penetration
depth of 4-5  cm be achieved for each  acceptable sample.

     Although  the 2-cm specification  is arbitrary, it will  ensure that:

     •   Relatively recent sediments are sampled

     •   Adequate volumes of sediment can be  obtained readily for
          laboratory analyses

     •   Data from different  studies  (historical  or  ongoing) can
         be  compared validly.

Sampling depths  other than  2  cm may be appropriate for  specific purposes.
For example, when toxicity determinations are made for marine dredge disposal
permitting,  sediments collected from depths  as great  as several meters
may be tested.  It should be remembered, however, that  if a sampling depth
other  than  2  cm  is used,  comparisons with data from 2-cm deep samples may
be questionable.
                                  12

-------
                                              Laboratory Sediment Bioassays
                                                       Sediment Collection
                                                                  May 1986
OPERATION OF  SAMPLER

     The sampling  device should be attached  to  the  hydrowire using a ball-
bearing swivel.   The  swivel will minimize the  twisting  forces on the sampler
during  deployment and  ensure that proper contact  is  made with the bottom.
For safety, the  hydrowire,  swivel,  and all  shackles should have  a  load
capacity at least three  times greater than the weight of a full sampler.

     The sampler should be lowered through the water column at a controlled
speed of approximately  1 ft per sec.  Under no circumstances should the sampler
be allowed to "free fall"  to the bottom, as this may result in premature
triggering, an excessive bow wake, or  improper orientation upon contact
with the bottom.  The  sampler should  contact  the  bottom gently, and only
its weight or piston  mechanism  should be used  to  force  it into the sediment.

     After the sediment  sample  is taken, the sampler  should be raised slowly
off the bottom and  then  retrieved at  a  controlled speed  of approximately
1 ft per sec.   Before  the  sampler  breaks the  water surface, the survey
vessel  should head  into  the waves (if present) to minimize  vessel  rolling.
This maneuver will  minimize  swinging  of the sampler after it breaks the
water surface.   If excessive  swinging occurs or  if  the  sampler strikes
the vessel during  retrieval, extra attention should  be paid to evaluating
sample disturbance  when  judging sample acceptability.

     The  sampler should be secured immediately after it is brought on board
the survey vessel.  If the sampler tips or slides around  before being secured,
extra attention should be  paid  to evaluating sample  disturbance.

SAMPLE ACCEPTABILITY CRITERIA

     After  the sampler is secured on  deck,  the  sediment sample  should be
inspected carefully before being  accepted.  The  following acceptability
criteria should be  satisfied:

     •    The sampler is not overfilled  with sample  so  that the sediment
          surface is pressed against the top of the  sampler

     •   ' Overlying water  is present  (indicates minimal leakage)

     •    The  overlying water is  not  excessively  turbid (indicates
          minimal sample disturbance)

     •    The  sediment  surface  is  relatively flat (indicates minimal
          disturbance or winnowing)

     t    The  desired  penetration  depth is achieved  (i.e.,  4-5  cm
          for a 2-cm deep  surficial sample).

If a sample does not meet  any  one of  these  criteria, it should be  rejected.

                                   13

-------
                                             Laboratory Sediment Bioassays
                                                       Sediment Collection
                                                                 May 1986
SAMPLE COLLECTION

     After a sample is judged acceptable, the following observations should
be noted on the field log sheet:

     •    Station location

     •    Depth

     •    Gross characteristics of the  surficial sediment

              Texture
              Color
              Biological structures (e.g., shells, tubes, macrophytes)
              Presence of debris (e.g., wood chips, wood fibers, human
              artifacts)
              Presence of oily sheen
              Odor (e.g., hydrogen sulfide, oil, creosote)

     •    Gross characteristics of the  vertical profile

              Changes in sediment characteristics
              Presence and  depth of  redox potential  discontinuity
              (rpd) layer

     •    Penetration depth

     t    Comments relative to sample quality

              Leakage
              Winnowing
              Disturbance.

     Before subsamples of the surficial  sediments are taken,  the  overlying
water must be removed.   The  preferred method  of removing this  water is
by slowly siphoning it  off  near one side of  the sampler.  Methods  such
as decanting the water or slightly cracking  the  grab to  let the  water run
out  are not  recommended,  as they may result  in  unacceptable disturbance
or loss of fine-grained surficial sediment and organic matter.

     Once the overlying water  has been  removed, the  surficial sediment
can be subsampled.  It  is  recommended that  subsamples be taken using a
flat  scoop  shaped like  a  coal  shovel.  The shoulders of the scoop should
be 2 cm high.  This device will allow  a relatively large  subsample to be
taken  accurately to a depth of 2 cm.   Coring devices are not recommended
because generally they  collect small  amounts  of surficial sediment, and
therefore require repeated  extractions to obtain a sufficient  volume of
material for analysis of  conventional sediment  variables.   A  curved scoop

                                   14

-------
                                             Laboratory Sediment Bioassays
                                                       Sediment Collection
                                                                 May 1986


is not  recommended because  it  does not sample  a uniform depth.  Because
accurate and consistent  subsampling requires  practice,  it  is  advisable
that an  experienced person perform this  task.

     Finally, sample  contamination during  collection must  be  avoided.
All sampling equipment  (e.g.,  scoops, containers)  should  be  made  of non-
contaminating material, and  should be cleaned  appropriately before use.
It is recommended that  all  objects coming  in contact with the  sample be
made of glass, stainless  steel, or TFE (i.e., tetrafluoroethylene; e.g.,
Teflon).  To avoid contamination, all  sampling equipment should  be  washed
thoroughly  with pesticide-grade methylene  chloride prior to initial use
and between  use for each sample.

SAMPLE HOMOGENIZATION

     Both compositing  of samples  for  chemistry and bioassay testing, and
subsampling  bioassay samples for different types of testing  require thorough
mixing (homogenization) of the initial  sample.  This is particularly important
when the contents of several  samples are required to  provide sufficient
material for testing.

     Compositing  and  homogenization  can be accomplished by transferring
sediment to  a solvent-rinsed glass or stainless  steel bowl  and thoroughly
homogenizing by stirring  with  a stainless  steel spoon and  spatula until
textural and color homogeneity are achieved.  The bowl and  all  utensils
should  be solvent-rinsed between composites,  and kept covered  with aluminum
foil to prevent airborne or other contamination.

CONCURRENT COLLECTION OF SEDIMENT CHEMISTRY  AND BIOASSAY SAMPLES

     If  sediment  chemistry samples are being collected along  with  sediment
bioassay  samples they should be collected  from the same sample.  Specifically,
composites  of several  samples  or of individual samples should be  combined
and mixed before aliquots are removed for sediment chemistry  and  sediment
bioassay determinations.  Sample homogenization and collection of  bioassay
aliquots should be  conducted so that chemical  aliquots are not contaminated
in  the  process.  The  protocols for collecting  aliquots  for analyses of
organic  compounds  and metals should be consulted in the appropriate  chapters
of  this notebook  to ensure that collection  procedures for bioassay  aliquots
are compatible with those recommended for chemical  aliquots.
                                  15

-------
                                              Laboratory  Sediment Bioassays
                                                             General QA/QC
                                                                  May 1986
           GENERAL QUALITY ASSURANCE/QUALITY CONTROL GUIDELINES


     The following  general  QA/QC procedures apply to all  sediment  bioassays.
QA/QC procedures  specific  to each bioassay for  which protocols  are detailed,
and  that  differ from  the following  generic  guidelines,  are incorporated
into individual  bioassay protocols.

NEGATIVE CONTROLS

     All bioassays  must be conducted using well-established  negative  (clean)
controls.   For every  test  series with a particular  organism,  one bioassay
test chamber or series of chambers must contain clean,  inert  material  plus
diluent seawater.  The  complete bioassay series  must be repeated  if more
than 10 percent of  the  control animals die  or show evidence of sublethal
effects.

POSITIVE CONTROLS

     All  bioassays should be  conducted  using  well-established positive
(toxic) controls.  These controls involve the  use of reference  toxicants.
Reference toxicants are used to provide insight into mortalities  or increased
sensitivity that may  occur as a result of disease,  changes  in  tolerance/sensi-
tivity, and   loading density.  Reference toxicants can also provide  insight
into nonlethal  effects  that  occur due to acclimation, insensitivity, or  stress
tolerance developed in  handling and bioassay.  Accordingly, concurrent bioassays
using a reference toxicant should be implemented  for each test  series.

TEST ORGANISMS

     Only  healthy  organisms  of similar size and life history stage  should
be used in bioassays.   Taxonomic  identifications of  bioassay  organisms
must be confirmed by  a  qualified taxonomist.

REFERENCE TEST SAMPLES

     Control  sediments generally are  those  from which test animals  were
collected.  As such,  physical and chemical sediment characteristics (e.g.,
grain  size,  organic content) may be very different from those  of  the  test
sediments.  Where this  is  the case, one or more reference sediments should
be  added  to   the test  series.  Reference sediments should be collected  from
an area documented to be  free from chemical contamination  and should represent
the  range of important  physical  and chemical  variables found  in  the  test
sediments.  Data derived  from such a sample, if in fact  it  is  totally  free
of  contamination, can be used to partition toxicant effects from unrelated
effects such  as those of  sediment grain size.

                                   16

-------
                                              Laboratory Sediment  Bioassays
                                                              General QA/QC
                                                                  May  1986
BLIND TESTING

     All treatment  and  bioassay containers should be randomized  and  testing
should be conducted  without  laboratory personnel knowing sample  identities.
Replicates of each treatment  should be assigned a code number during  testing
and randomized in the  test sequence.

MAINTENANCE/MEASUREMENT  OF WATER QUALITY

     Bioassays  involving exposure of organisms in  aqueous media  require
that the media be uncontaminated and that proper  water quality  conditions
be maintained to ensure the survival of the organisms, and to  ensure  that
undue stress is not exerted on the organisms unrelated to the test sediments.
At a  minimum, the  following variables  must  be  measured at the beginning
and termination of testing:   salinity, dissolved oxygen, pH, and  temperature.

STANDARD LABORATORY  PROCEDURES

     Standard laboratory procedures must be followed in all testing. These
include proper documentation, proper cleaning, avoidance of contamination,
and  maintenance of appropriate test conditions.  All unusual  observations
or deviations from established procedures must be recorded and reported.
                                   17

-------
                                             Laboratory Sediment  Bioassays
                                                         Amphi pod  Bioassay
                                                                  May 1986
                        AMPHIPOD SEDIMENT BIOASSAY
USE AND LIMITATIONS
     The amphipod  (Rhepoxynius abronius)  sediment bioassay is used  to charac-
terize the toxicity of marine sediments.   This bioassay may  be  used alone
as a  screening tool in broad-scale sediment surveys, in combination with
sediment chemistry and in situ biological  indices, and in laboratory  experi-
ments addressing a variety of sediment and water quality manipulations.

     The  amphipod bioassay  is  not  appropriate for sediments that  have  an
interstitial salinity of  less than  15  ppt.   In addition,  the following
constraints apply:

     •    An interstitial  salinity  of  25 ppt  is necessary to  ensure
          that there are no salinity effects.  If interstitial salinities
          are between 15 and 24 ppt,  they must be adjusted upward.

     •    Grain size may have an effect on  the animals  at extremes
          of fine  and coarse material.    If  the  clay  content  of the
          test  sediments exceeds 50 percent or  the  gravel content
          exceeds  35 percent,  controls   for  the  effects of particle
          size  distribution are recommended.

     •    As in  all  bioassays  using natural populations, there  is
          a possibility that relative sensitivity  of the  amphipods
          will  vary with season or other factors.   Accordingly,  a
          positive control  is  recommended.  This  should comprise  a
          96-h LC50 measurement with a reference  toxicant (e.g., CdCl2
          or NaPCP) conducted in the  absence of sediment.

     t    Identification of R. abronius must be confirmed by a qualified
          taxonomist  prior to initiation  of the bioassay, and represen-
          tative  specimens should be preserved and archived  for future
          reference.

     •    Predators  generally are  not  a problem  in the Dioassay, but
          potential problems can be avoided by observation and reasonable
          care.
                                  18

-------
                                              Laboratory Sediment  Bioassays
                                                          Amphi pod  Bioassay
                                                                   May  1986
FIELD PROCEDURES

Collection

Test Animals--

     Amphipods  are obtained  using benthic  grabs (e.g.,  van  Veen,  Smith-
Mclntyre) or small  dredges.  If  a dredge is used, a short haul  (10 m) will
minimize potential damage  to the  animals during collection.  Approximately
one third more animals  than  are  required for  the  bioassay are collected.
Surface and bottom  seawater salinity and temperature are measured  at  the
collection site.  Sediment  temperature  is recorded from the first and last
dredge sample.  It  is recommended that  bioassays be conducted  within  10 days
of amphipod  collection.

Sediment--

     Both control  and  test sediment should be collected in solvent-cleaned
glass jars having TFE-lined lids.   Each jar  should  be filled completely
to exclude  air.  A minimum  sediment sample size of 0.25 L  for each bioassay
beaker is recommended  for both kinds of sediment.

Processing

Test Animals--

     Contents of the dredge  are  gently  washed  into a container using seawater
of similar temperature  and salinity  to that at  the collection depth.  Samples
that show evidence of contamination (e.g., oil sheen)  are rejected. Amphipods
will typically bury in  the  sediment and if  necessary can be  held  in  the
containers  for  several hours (at  the  bottom  temperature)  prior to sieving.
It is preferable to minimize the  delay from collection  to sieving.   To
avoid handling stress,  each  dredge  sample is placed in a separate container.
Animals are maintained  and  transported  in clean coolers (with  ice as  neces-
sary).

Sediment--

     Both control  and  test sediment  should  be stored at  40 c  in the dark.
Holding time should not exceed 14 days.

LABORATORY PROCEDURES

     The  laboratory procedures are those described by Swartz et al.  (1985)
with the following changes  incorporated:

     t    The salinity  of the  overlying water  is increased  to 28 ppt

     •    Reburial  is omitted  as a  response variable

                                   19

-------
                                              Laboratory Sediment Bioassays
                                                         Amphipod Bloassay
                                                                  May 1986
     •    A 1.0-mm screen  is  used to sieve  out  amphipods  prior  to
          initiation of  testing

     t    Holding  time  for amphipods is standardized to between 4 and
          10 days

     •    Sediment holding time prior  to testing  is set at a maximum
          of 14 days

     •    Additional  details are  provided concerning maintenance and
          transportation  of  amphipods, confirmation of taxonomic identi-
          fications,  and  freeing  of  amphipods  trapped during testing
          by water  surface  tension

     •    A specific  procedure for  interstitial   salinity adjustment
          is recommended

     •    A positive control is stipulated and  data  on 96-h LC50 values
          for  two reference  toxicants are provided.

Test Animals

Sieving—

     A 1.0-mm  sieve  is  used to  remove j?. abronius from sediment.   Mature
amphipods (3.0-5.0 mm total length)  are used in  the sediment bioassay.
Gentle  sieving is  essential to reduce handling stress.  The sieve is placed
in a large tub filled with seawater at ambient collection site bottom salinity
and  temperature.  The  entire contents of each holding container, including
water, are washed through the sieve using seawater  pumped  at  low pressure
through  a fan  spray nozzle.  The sieve can be  shaken gently, but the bottom
of the screen must  be  beneath the  water  surface at all  times.  Material
retained on the screen  is washed into  buckets for sorting.   Large pieces
of detritus and obvious  predators are  discarded.   If there is a delay of
more  than 1 n  before sorting  begins,  the buckets should contain enough
sieved sediment to  allow the amphipods  to bury.  The buckets  must be kept
at collection  site  temperature.   Aeration may be necessary.

Sorting--

     An  aliquot  of detritus or  sediment containing  amphipods  is  placed
in a sorting tray.  Healthy, active animals are removed with a  bulb pipette
(5-mm opening)  and placed  in 10-cm diameter finger  bowls filled with 28 ppt
seawater and a  2-on deep layer of 0.5-mm sieved collection  site sediment.
Twenty  amphipods are held in  each bowl  and  enough bowls are prepared to
provide at least one third more specimens  than are required for  the bioassay.
For R.  abronius, seawater temperature during sorting must not  exceed  18° c.
Filled finger bowls are  submerged in  holding  tanks supplied  with flowing

                                  20

-------
                                             Laboratory Sediment  Bioassays
                                                         Amphipod  Bioassay
                                                                  May  1986


water or aeration where temperature and  salinity approximate bioassay  condi-
tions.  If temperature and salinity adjustments are necessary,  they should
be made  gradually.  Healthy  R.  abronius will  remain in the  finger  bowl
sediment and  can be retrieved easily when  the  bioassay is set up.   Amphipods
should  be acclimated to  laboratory conditions for a minimum of 4 days and
a maximum of  10 days  before testing.

     The  identification of amphipods  as R. abronius must be confirmed  by
a qualified taxonomist.  In addition, representative specimens from each
bioassay series should be preserved and  archived for future reference.

Control Sediment

     j*. aoronius  typically inhabits well-sorted, fine sand.  Suggestions
for sTeving and settling may have to be  adjusted for other sediment types.

     Approximately  0.25 L of control sediment  is collected for each bioassay
beaker.  This sediment is sieved twice:  first, to remove  the test species
and  other macrobenthos and  second, to adjust interstitial water salinity.
The entire contents of one or more  sediment  samples,  including water and
suspended particulate matter, are  sieved  through a 0.5-mm screen without
allowing overflow  from the container.  After  the first sieving, the sediment
is allowed to settle for  at  least  4  h.  Overlying water is then decanted
and the sediment resieved through a 0.5-mm screen  into water of the bioassay
salinity (28 ppt for  R. abronius).  Again, the sediment is allowed to  settle
for at least 4 h (preTerably overnight), overlying water  is  decanted, and
the control sediment  is held at 40 c until  the bioassay chambers are prepared.

Test Sediment

     Approximately 0.25 L of test sediment  are necessary for each bioassay
beaker.  The natural  geochemical  properties of test  sediment collected
from  the field must  be within the  tolerance limits of the test species.
R. abronius may be  adversely affected by salinity  stress if the interstitial
water salinity is below  25 ppt.  Interstitial  salinities below 25 ppt may
therefore require  adjustment upwards.  This species  is  tolerant of a broad
range of sediment types.   However, controls for the effects of particle
size distribution  are recommended  if  the  clay content  exceeds 50 percent
or the gravel content exceeds 35 percent.

     The  j*.  abronius test requires a minimum water column  salinity regime
of 28 ppt7  When the interstitial  (i.e., pore water)  salinity  is below
25 ppt, it must be raised  if this test is to  be used.   The following procedure
is recommended in  such cases.  The  interstitial salinity  of the sediments
is determined (e.g., by  refractometer)  and the  sediments are placed  in
the bioassay chambers with overlying  water of a  salinity calculated  to
raise interstitial  salinities to  a minimum of 25 ppt. The sediments are
then  carefully and  slowly stirred by  hand  with a  clean glass rod for 1
min,  and allowed  to settle and  equilibrate.   The majority  (approximately

                                   21

-------
                                              Laboratory Sediment Bioassays
                                                         Amphi pod Bioassay
                                                                  May 1986


75 percent) of the overlying water is then carefully decanted and the inter-
stitial salinity in each  chamber confirmed prior  to bioassay initiation.
The  decant water  is  retained, salinity adjusted  if  necessary, and used
as the overlying water in  the bioassay.  Sediments  are  mixed after  adding
the decant water.  After  the bioassays terminate,  the interstitial salinities
are reconfirmed.

Bioassay Seawater

     Seawater  used in the bioassay is maintained  at a salinity of 28 +1 ppt
and  temperature  of  15 +_lo  c.   If a series of experiments is planned, test
temperature and salinity should be  the same throughout  the series.   The
bioassay  seawater must  be uncontaminated, which  may necessitate collection
of seawater at  the amphipod collection site.  The quantity of seawater
required  is dependent on sieving  and  holding needs  and on the number of
bioassay chambers.

     Bioassay  seawater is  passed through a filter  with 0.45-um pore diameter.
If necessary,  salinity is  reduced by addition of deionized  distilled water
or raised  by addition  of  clean oceanic water or reagent  grade chemicals  (ASTM
Practice for Conducting Acute Toxicity Tests with Fishes, Macro invertebrates,
and Amphibians [E 729]).   Seawater is prepared within 2  days of the bioassay
and stored in  covered, clean containers at the bioassay  temperature.

Facilities and Equipment

     The bioassay chamber is  a  standard 1-L  glass beaker (10-cm internal
diameter)  covered with an  11.4-on diameter glass watchglass.   The beakers
are  placed in a  shallow water bath  or temperature-controlled room with
overhead aeration  source.  Aeration  to each beaker is  provided through
a 1-mL  glass  pipette that  extends between the beaker  spout and watchglass
to a depth not closer  than 2 cm from the sediment  surface.   Air is bubbled
into the  beakers at a rate that does not disturb  the sediment. The bioassay
temperature is maintained by either  the water bath or  room temperature
control.

     All glassware is cleaned by washing with laboratory detergent; rinsing
in turn with  distilled water, 10 percent nitric acid (HN03) or hydrochloric
acid (HC1), and  distilled water.   Large plastic  containers and plastic
sieves  used for preparation and storage of sediment  and  seawater are precon-
ditioned initially by  soaking for 24  h  in seawater  and rinsed after  each
use with  clean seawater.   They are used  only for bioassays and stored  in
a clean room.   Sieves  and containers used to collect and  store amphipods,
seawater,  and control  sediment are kept separate  from those used for test
sediment.
                                   22

-------
                                             Laboratory  Sediment Bioassays
                                                        Amphipod Bioassay
                                                                 May 1986
Bioassay  Procedure
     Approximately  175  ml of test sediment are placed  in  the  bottom of
the 1-L bioassay chamber to create  a 2-on deep  layer.   Beakers are  filled
to 950 ml with  28 ppt seawater,  covered with a watchglass, and placed in
a 150 c water  bath.  Constant illumination  is provided  by overhead  lights.
Water in  the  beakers is aerated without disturbing  the  sediment surface.
Twenty amphipods  are placed  in  each beaker. The  bioassay is terminated
after 10  days of exposure.

     The primary  response criterion  is  survival after  10  days  exposure
to test or control sediment.   The secondary response criterion is daily
emergence of  amphipods from toxic  sediment.

Initiation--

     The day before the bioassay is  initiated, each test sediment sample
is mixed  within its  storage  container and  an  aliquot sufficient  to make
a 2-cm deep  layer  is added  to  a bioassay  beaker.   For replicate bioassay
samples, the weight of sediment necessary to make a 2-cm deep layer (approxi-
mately 175 ml)  in  the  first  beaker  is added to the other replicates.  The
same  procedure applies to control  sediment.  Treatments are randomly assigned
to prenumbered  bioassay beakers.

     The sediment  aliquot in the beaker  is settled by  smoothing with a
spoon, and bubbles are removed by tapping the  beaker against  the  palm of
the hand.  A  disk  (attached to a string for removal)  is placed on the sediment
surface.  This minimizes sediment disruption  as bioassay  seawater  is added
up to  the 750-mL mark  on the beakers.   This disk  is  removed and rinsed
in bioassay  water between beakers and changed  between  treatments.   The
beakers  are  covered with watchglasses, put  into  the  150 c water  bath or
temperature-controlled room, and aerated.  The beakers are allowed to  equili-
brate  overnight to  bioassay conditions.  Normal room lighting is maintained
continuously during the bioassay.   If the  experimental  design requires
monitoring  of  sediment  chemistry (e.g., metals,  total  volatile  solids,
Eh), additional beakers must be set up for  this  purpose.  Monitoring the
quality of seawater  overlying the sediment can be accomplished in the bioassay
beakers without disturbing  the sediment.   Temperature  is  recorded from
a  thermometer  maintained  in  a separate beaker containing control sediment
and bioassay  water but no amphipods.

     On the  day the bioassay is initiated,  amphipods are distributed  among
all beakers  so  that  each receives 20  individuals.  It is  usually not  logis-
tically  possible to distribute amphipods  to all beakers at the same  time,
so it necessary  to select a portion of the  beakers  (as many as  15) to be
processed  together.  The exact number of  beakers to receive  amphipods at
one time is  dependent on the size and  design of the experiment.   At least
one  replicate  from  each treatment, including control sediment,  is processed
at a time if  possible.  Otherwise, selection  is random.

                                  23

-------
                                             Laboratory Sediment  Bloassays
                                                         Amphi pod  Bioassay
                                                                  May 1986
     Amphipods are  removed from the  holding sediment using a 1.0-mm  sieve,
and then  transferred to sorting trays.  About  one third more  fingerbowls
are removed  from the holding tank at one  time than are required for the
number of beakers.  This allows selection  of active, healthy  animals for
the bioassay.  Amphipods are removed  from  the  sorting tray and  sequentially
distributed among clean 10-cm fingerbowls each  containing 150 ml of bioassay
seawater  without sediment.   The number of amphipods distributed to each
fingerbowl  is  recounted by transferring  them  to a  separate fingerbowl.

     Amphipods are added to the bioassay beakers  by placing a black plastic
disk on the  seawater  surface  and gently  pouring the entire  contents of
the fingerbowl  into  the beaker.  The fingerbowl  is washed with bioassay
water to  remove adhering amphipods.  The seawater level  is  brought  up to
950 mL with  bioassay water,  and the disk  is removed and  rinsed between
samples.   Amphipods are allowed to bury  in  the sediment  and any that are
floating  on the seawater surface are pushed down with the edge of the  beaker
cover.  After  1 h, amphipods that have not  buried  are removed and replaced.
Normally, less than 1  percent of the animals  will  fail to bury in 1 h.

Monitoring--

     If samples  for  chemical  analysis are desired, seawater  and sediment
samples can be taken from  beakers at the  initiation of  the bioassay.  A
small  quantity  of seawater  can be taken  from beakers at  the initiation
of the bioassay,  but  chemistry beakers have  to be sacrificed to  obtain
sediment samples.  This is accomplished by siphoning the overlying seawater
without disturbing the sediment surface  and then taking appropriate sediment
aliquots for  chemical analyses.  It  is not  necessary to add  amphipods to
chemistry beakers that are sacrificed at the initiation  of  the bioassay,
but amphipods are added to  those sacrificed  later.  Certain  sediment and
water quality  variables (e.g., dissolved oxygen, pH, Eh)  can be monitored
by inserting analytical probes into the  chemistry  beakers.

     During the course of the bioassay,  certain observations are made  daily.
Temperature  in the beaker set up for this purpose is monitored.  Lighting
and aeration  systems are  checked.  Each beaker is carefully  examined but
not disturbed  except  for the  temporary removal of the  aeration  pipette
and watchglass.   Notes are made on sediment appearance and unusual conditions.
The number of  amphipods that have emerged from  the sediment, either floating
on the water  surface or lying on top of the  sediment, is counted and recorded.
Amphipods that have emerged are not removed,  even  if they  are dead.   These
data  are used to document the temporal  pattern of emergence.  Amphipods
trapped by surface tension at the water  surface are gently  pushed down
with a clean  instrument (e.g., pipet, glass rod,  beaker cover).
                                     24

-------
                                              Laboratory Sediment  Bioassays
                                                          Amphipod  Bloassay
                                                                   May 1986
Termination--
     The bioassay is terminated after 10  days of exposure.   After daily
observations are recorded, the contents of the bioassay beakers  are sieved
through  a  0.5-mm  screen.   Material  retained  on  the screen is  placed  in
clean bioasssay water in a sorting  tray.  The number of live and  dead  amphipods
is recorded.  The  sum  of  these numbers  may not  always equal 20 because
of death and subsequent  decomposition of amphipods.   An amphipod  is counted
as alive if there  is any sign  of life (e.g., pleopod  twitch  observed under
magnification,  response  to gentle prodding with a clean instrument).

Experimental Design

Logistics--

     A typical  sediment bioassay involves about 50 to 60 bioassay beakers.
Collection  and  preparation of  animals, sediment, and  seawater  requires
at least four people  for  2 days.   Three or four people are  required  on
the days experiments are initiated and terminated.  One person can monitor
the experiment in progress.

Controls--

     Five  replicates of the  amphipod collection  site control  sediment are
included  in  all  bioassays.  These beakers comprise a negative (clean)  control
that allows comparisons among  experiments and  among laboratories of the
validity of the procedures  used in  individual  investigations.  At least
90 percent of the  amphipods  in  the control replicates should survive.  The
design of field surveys may include a reference sediment  involving five
replicate samples from  an area believed to be free from sediment contamination.
This provides a site-specific  basis  for comparison of potentially toxic and
nontoxic conditions.  Experiments in which contaminants are added to sediment
may require control replicates to determine effects of solvent addition.

     A positive (contaminated) control  is also  required for all testing.
This involves determining 96-h LC50  values for R. abronius exposed  in clean,
filtered seawater without sediment to  reference  toxicants (following standard
bioassay  procedures and under  the  same general test conditions as the sediment
bioassays).  Such  data are necessary to determine the relative  sensitivity
of the animals (e.g., seasonal  difference  in  sensitivity)  for each test
series  to  ensure  comparability of the data.  Two commonly used  reference
toxicants are reagent-grade  CdCl2 and  NaPCP.  Swartz et al. (1986)  determined
a 96-h LC50 of 1.61 mg/L for CdCl2,  while Cummins and Gangmark  (in preparation)
found that  the 96-h LC50 for  NaPCP  ranged  from 0.30  to  0.39 mg/L over  a
2-yr period.  Either reference toxicant may be used, but the acute  lethality
results must be reported  along  with the sediment  bioassay results.  Bioassays
to establish an  LC50  involve four  or five  logarithmic concentration series
and a control.   At least one treatment should give  a partial response below
                                   25

-------
                                              Laboratory Sediment  Bioassays
                                                          Amphipod  Bioassay
                                                                   May  1986


the LC50 and one above the  LC50.  Statistical procedures for the LC50 estimate
are given in APHA (1985).

Response Criteria--

     Emergence  from  sediment and survival  are the  two response  criteria
examined in all  bioassays.   Emergence data are used to monitor" a behavioral
response of the amphipods during  the 10-day exposure.  Survival  after  10
days of exposure is the primary  criterion of toxicity.  Each of these  response
criteria must be monitored  in  a  "blind" fashion;  that is,  the  observer
must have no knowledge of the  treatment of  the  sediment in  the  beakers.
This is accomplished through randomization of beaker numbers.

DATA REPORTING REQUIREMENTS

     The  following data should be reported by all laboratories performing
this bioassay:

     •    Water quality measurements during testing (i.e.,  DO, temperature,
          salinity,  pH).

     •    Daily emergence for each beaker and the  10-day mean  and
          standard deviation for each- treatment

     •    10-day  survival   in  each beaker  and  the mean  and  standard
          deviation  for each treatment

     •    Interstitial salinity values  of test sediments

     •    96-h LC50  values  with reference toxicants

     •    Any problems that may have  influenced data quality.
                                   26

-------
                                             Laboratory Sediment  Bioassays
                                                   Bivalve Larvae  Bioassay
                                                                  May 1986
                     BIVALVE  LARVAE SEDIMENT BIOASSAY
USE AND LIMITATIONS

     The bivalve  larvae bioassay technique  is described in Standard  Methods
(APHA 1985)  and  by ASTM (1985) as a rapid  and reliable indicator of environ-
mental  quality.  Pacific oysters (Crassostrea gigas) and blue mussels  (Mytilus
edulis) are  recommended for testing.  During  the first 48 h  of embryonic
development,  fertilized oyster and mussel  eggs  normally develop into  free-
swimming, fully shelled  larvae (prodissoconch I).  Failure of the eggs
to  survive  or the proportion of larvae developing in an abnormal manner
is used as an  indicator of toxicity.

     This sediment  bioassay  can  be used to characterize the toxicity  of
marine sediments.  It may be used alone  as a screening tool  in broad-scale
sediment surveys,  in combination with sediment chemistry and in  situ biological
indices, and in  laboratory  experiments  addressing a variety of sediment
and water quality  manipulations.

     The bivalve larvae bioassay can be  used in  sediments that have interstitial
salinities less  than 1 ppt, as the sediments are  mixed and equilibrated with
seawater prior  to testing.  However, because further testing  is required  to
determine the validity of using this technique with  such low salinity sediments,
this bioassay is not recommended  for  sediments that have an interstitial
salinity of  less than 10 ppt.  In addition,  the  following caveats apply:

     •    Bivalve larvae  such as those  of C^.  gigas normally are not
          associated with the  types of sediments  that generally  are
          tested  using this  method.  Hence,  this bioassay  is  only
          an  indicator of  relative  toxicity  and is not  usually  of
          direct ecological significance.

     •    Spawning of £.  gigas occurs  naturally in the  Puget Sound
          area in  summer.  The natural spawning  period for H.  edu 1 is
          is  late spring to  early summer.   Both  of these  bivalves
          can be  induced  to spawn  at  other  times of the year, but
          may show decreased viability  of gametes.  Gamete viability
          may  also vary depending on the brood  stock  used.  Accordingly,
          a  positive  control is  recommmended.  This should comprise
          a  48-h EC50 measurement  with a reference toxicant  (either
          CdCl£  or NaPCP) conducted  in the absence of sediment.

     •    It  is possible  that abnormalities  induced during testing
          may  be underestimated due to poor  recovery of living  larvae
          from  the sediments.  Accordingly,  it is recommended that

                                   27

-------
                                              Laboratory Sediment Bioassays
                                                   Bivalve Larvae Bioassay
                                                                  May 1986


          a few sediment  samples from each  set  of bioassays conducted
          with  this  technique be examined  to  determine whether  living
          larvae  are present  in the sediment.   The results  should
          be  quantified and reported.

FIELD PROCEDURES

Collection

     Both  control  and test sediment  should  be collected in solvent-cleaned
glass jars having TFE-lined lids.  Each  jar should  be  filled completely
with  sediment  to exclude  air.  A minimum sediment sample size of 20 g for
each bioassay chamber  is recommended for both kinds of sediment.

Processing

     Both  control  and test sediment should be  stored at 40 c in the dark.
Holding time  should  not exceed 14 days.

LABORATORY PROCEDURES

     The  following  procedures  apply equally to  larvae of both £. gigas
and M. edulis.  and  are as described by Chapman and Morgan  (1983) with  the
following changes incorporated:

     •    The salinity of test water is adjusted  to 28 ppt

     •    Replication  is  increased from two to  five  to  al low adequate
          statistical  comparisons

     •    Larvae of  M. edulis are included in the bioassay protocol

     t    Sediment  resuspension  in the bioassay containers is adequately
          accomplished by  vigorous shaking  for 10  sec; there  is no
          need  to rotate the containers for 3 h at  10  rpm

     t    Twenty grams  of sediment  is  suspended in  1  L  of seawater
          rather than  15 g  in 750 mL

     •    Sediment  holding time prior  to testing is set at a maximum
          of  14 days

     •    pH  is not adjusted before  the bioassay starts and is only
          monitored

     •    A positive control is  recommended

     t    Additional  details provided by ASTM (1985) for conditioning
          and spawning adults are included.

                                  28

-------
                                              Laboratory Sediment Bioassays
                                                    Bivalve Larvae Bioassay
                                                                  May 1986
Bioassay Species

     For a given  test or series  of related  tests, adult bivalves (brood
stock) are obtained  from the same source, either commercial  rearing facilities
(oysters)  or a chemically  uncontaminated area (mussels).  Within 24 h of
collection or purchase,  adults should be transported  to  the test  laboratory
and  placed into  flowing  seawater  similar in character  to that from which
they were taken.   Rough handling;  extended periods  of desiccation;  and
abrupt  changes in  temperature, salinity, or other  water quality variables
must be avoided as these may  induce premature spawning  or render the  stock
useless for later controlled spawning or both.  Upon receipt, adults should
be cleaned of fouling  organisms and detritus and placed  in flowing  seawater
for conditioning.

     Adult bivalves are held  at  recommended conditioning temperatures to
stimulate final maturation of the gametes.  The desired conditioning  temperature
(20 +20 c for oysters  and 14  +20 C for mussels) should  be attained gradually
at an  increment  not  exceeding 2° C per day.  Conditioning may  extend from
a few days to several  weeks depending on the physiological and gametogenic
status  of the adults.  The  length of the conditioning  period  is determined
empirically by periodic sacrifical examination and spawning of representative
individuals.  Adults  should be spawned  or discarded within 2-3 wk after
attaining acceptable maturity  because gamete quality will deteriorate rapidly
with excessive conditioning.

Bioassay Sediment

     The bivalve  larvae bioassay is conducted with  sediment  controls in
addition to  seawater  controls.   Control  sediment  typically consists of
material collected from an area documented to be free of chemical  contamination
and of  proven nontoxicity to  bivalve larvae.

Bioassay Seawater

     Seawater used in the bioassay  is maintained at a salinity of  28 +1 ppt
and  temperature  of 20 +10 C.  The  bioassay seawater must be uncontaminated
and of acceptably low toxicity.  The standard biological  criterion  of accepta-
bility is that the larvae, spawned  by adults in the dilution  water, must
not  incur more than 10 percent abnormal development or 30 percent mortality
during  48 h of exposure to the dilution water  (ASTM 1985).

     If necessary, salinity of the bioassay water is reduced  by addition
of deionized distilled water  or raised by  addition of clean oceanic water
or  reagent grade chemicals (ASTM  Practice  for  Conducting Acute Toxicity
Tests with Fishes, Macroinvertebrates, and  Amphibians  [E729]).   Seawater
is  prepared within 2  days of use  and  is stored  in clean, covered containers
at the  requisite temperature.


                                  29

-------
                                             Laboratory Sediment Bioassays
                                                   Bivalve  Larvae Bioassay
                                                                 May 1986
Facilities  and  Equipment
     The bioassay  chamber is a clean,  acid-rinsed, 1-L glass  bottle with
a screw-top  lid.  Bioassays are conducted with the bottles in  shallow water
baths,  incubators,  or temperature-controlled rooms.

     If adults are  to be conditioned  for  spawning out-of-season,  a continuous
supply  of temperature-controlled seawater  is  needed.  Raw  seawater can
be used  for holding and  conditioning, but  feeding the adults a  natural
or cultivated  alga  is necessary to deter  starvation.  The flow  rates used
for adult conditioning must be high enough to prevent water quality degradation
and provide  as much  food as possible  to the adults.

     Tanks and  trays  are necessary  for holding the  adults, and  a water
bath, incubator, or  temperature-controlled  room is necessary  during the
bioassay.   Adult  holding and conditioning tanks should be cleaned at least
weekly to prevent accumulation of organic matter and bacteria.   Dead specimens
should  be removed   immediately  and  the tanks  cleaned.  The tanks  should
be cleaned with  detergent and rinsed  with clean seawater, and  if microbial
growth  is present, rinsed with  200  mg/L of hypochlorite and  then seawater.
With enriched  waters and elevated conditioning temperatures, more  frequent
cleaning may  be required.   All  new test chambers should be  washed with
detergent and  rinsed in  turn  with water, pesticide-free acetone,  water,
acid, and then water.

Bioassay Procedure

     The bioassay  procedure can be  summarized as follows.  Adult bivalves,
conditioned as necessary in the laboratory, are  induced to spawn  with selected
thermal  and biological  (i.e.,  sperm) stimulation.  Selected densities of
the resulting embryos are exposed to the test or control  sediments for
48 h, during  which the  embryos normally will develop into prodissoconch  I
larvae.   Toxicity is measured based on abnormal shell development and larval
death.

     The bivalves  are spawned  by rapidly  raising  the  water temperature
to 5-100 c  above the conditioning  temperature.   Females are additionally
stimulated to  spawn  by the addition of sperm from a  sacrificed or naturally
spawned male.

     Spawning  is conducted  by  placing  the  bivalves  in individual, clean
Pyrex dishes containing filtered, UV-treated  seawater.   Fertilization is
accomplished  within 1 h  of  spawning by combining eggs and  sperm in a 1-L
Nalgene beaker.  The  fertilized eggs are then washed  through a  0.25-mm
Nitex  screen  to remove  excess  gonadal material  and suspended  in 2 L of
filtered, UV-treated seawater at incubating temperature.  The embryos are
kept suspended by  frequent  agitation using a perforated plunger and used
in the bioassay  within 2 h of fertilization.  When microscopic examination
of fertilized eggs reveals  the formation of polar bodies,  egg density is

                                  30

-------
                                              Laboratory  Sediment  Bioassays
                                                    Bivalve  Larvae Bioassay
                                                                  May 1986


determined  from triplicate counts of the number of eggs in 1.0-mL samples
of a 1:99 dilution  of  homogeneous egg suspension.

     Sediment  bioassays are  conducted  in  clean (i.e.,  rinsed with dilute
acid) 1-L glass bottles.   Twenty  grams (wet weight)  of the appropriate
sediment is  added  to each bottle and volume is brought up to 1 L with filtered,
UV-treated seawater  (28 ppt salinity)  to make  a  final  concentration in
all  containers of 20  g (wet weight)  of sediment  per liter of  seawater.
All test treatments  are  replicated five times.  The sediment controls contain
20 g/L of clean sediment.   In addition, a control  series is prepared  consisting
of clean seawater  without sediment.

     The sediments  are  suspended by vigorous shaking for 10 sec, the embryos
are added, and  the  suspended  sediments are allowed to settle.  No  additional
agitation is provided.   The  seawater control is treated  similarly.

     Within  2  h of fertilization, each container  is inoculated with 20,000-
40,000 developing  embryos to give  a  concentration  of about 20-40 per mL.
The  embryo  concentration at 0 h should be confirmed by  collecting  replicate
10-mL samples from control cultures and preserving them in 5-percent  formalin.
The  containers are covered  and incubated for 48  h at 20  +lo c under a  14-h
light:10-h dark  photoperiod.   Test vessels are not aerated  during the  bioassay.

     After  48  h,  the bioassay  can be terminated  in  one of two  ways.  In
the first method, the contents of each container are carefully poured through
a 38-um Nitex sieve without disturbing the settled sediment,  thereby retaining
and concentrating  the  surviving  larvae (larvae caught in the  sediments
are  almost  always dead).   The concentrated larvae are  washed into a 100-mL
graduated cylinder, quantitatively  transferred  to 8-mL screw-cap  vials,
and  preserved in  5 percent buffered  formalin.   In the  second termination
method, the water and  larvae  overlying the settled  sediment in each  container
are  carefully poured  into  a clean 1-L beaker.  This water   is then stirred,
and a 10-mL aliquot of the  well-mixed sample is removed  by pi pet  and  placed
in  a  10-mL  screw-cap  vial.  The  contents of each vial is preserved  in  5
percent buffered formalin.

     Preserved samples (equal in volume to those containing 300-500 larvae
in controls)  are examined in  Sedgewick-Rafter cells under 100X magnification.
Normal  and  abnormal  larvae are  enumerated  to  determine percent survival
and percent  abnormality.   Percent  survival  is  determined as  the  number
of larvae surviving in  each  test container relative to the seawater control.
Larvae that fail to  transform  to the fully shelled, straight-hinged,  'D1-shaped
prodissoconch I stage  are considered abnormal.

     Salinity  and dissolved  oxygen are initially adjusted in each  container
to 28 ppt and  a minimum of  90 percent saturation, respectively.   These
variables and pH are measured  in each container at the start  and  termination
of the bioassay.


                                   31

-------
                                              Laboratory Sediment Bioassays
                                                    Bivalve Larvae Bioassay
                                                                   May  1986
Controls

     Five replicates of  control sediment  and five  replicates of control
seawater alone  are  included  in  all  bioassays.   These comprise negative
(clean) controls that allow comparison among experiments and among laboratories
of the validity of the procedures used in individual  investigations.   At
least  70  percent of the  larvae must survive the 48-h exposure with seawater
alone, and of these  at least 90  percent must  show no abnormalities.   The
design of field surveys may include a reference sediment from an area believed
to be  free from chemical contamination.  This provides a basis for comparison
of potentially toxic and  nontoxic conditions.   Experiments in which contaminants
are added to sediment may require control replicates to determine effects
of solvent addition.

     A positive (toxic) control  is also required.  This involves determining
48-h LC50 and EC50 values for bivalve larvae exposed to reference toxicants
in clean, filtered,  UV-treated  seawater without sediment (following standard
bioassay procedures and under the  same general test conditions as the sediment
bioassays).  Such  data  are necessary to determine the relative sensitivity
of the larvae.  Two  commonly  used reference  toxicants are reagent-grade
CdCl2  and NaPCP.   Either reference  toxicant may be used, but the results
must be reported along with the  sediment bioassay results.  Bioassays to
establish an LC50 or an  EC50 involve four or  five logarithmic concentration
series and a control.  At least  one treatment  should give a partial response
below the LC50 and EC50 and one  above the LC50  and EC50.  Statistical procedures
for the LC50 and EC50 estimates  are given in APHA  (1985) and ASTM (1985).

DATA REPORTING REQUIREMENTS

     The  following   data  should be reported  by all laboratories performing
this bioassay:

     •    Water quality measurements at the beginning and end of testing
          (e.g., dissolved oxygen, temperature, salinity, pH)

     •    Individual replicate  and  mean and standard deviation data
          for larval survival after 48 h

     •    Individual replicate  and  mean and standard deviation data
          for larval abnormalities after 48 h

     •    48-h LC50  and EC50 values with reference toxicants

     t    Data on larval  presence in the sediment

     •    Any problems that may have  influenced data quality.
                                   32

-------
                                             Laboratory Sediment Bioassays
                                              Anaphase Aberration Bioassay
                                                                 May  1986
                  ANAPHASE ABERRATION SEDIMENT BIOASSAY
USE AND LIMITATIONS

     This sediment  bioassay is used to  characterize the genotoxicity  of
marine sediments.  It may be used alone as a screening  tool  in broad-scale
sediment surveys, in combination with sediment chemistry and in  situ biological
indices, and  in  laboratory  experiments addressing  a  variety of sediment
and water quality manipulations.

     This bioassay  can  be used with any type  of sediment, regardless  of
the interstitial salinity  or grain size characteristics.   However, the
following caveats apply:

     •    The bioassay depends on a chemical  extraction procedure
          that is specific for  neutral,  nonionic organic compounds.
          Other  classes of contaminants such  as metals and  highly
          acidic  and basic organic  compounds are not efficiently ex-
          tracted.   Thus,  characterization of  sediment toxicity is
          directed  towards  neutral compounds  such as aromatic  and
          chlorinated hydrocarbons.

     •    "Natural" genotoxicity  may  occur in  marine sediments due
          to  the  decomposition of plant  species  containing genotoxic
          substances  that evolved as  a  means  of protecting  plants
          from parasites  and  predators.  Thus,  positive  genotoxic
          responses may be  noted  in areas generally regarded as pris-
          tine.

FIELD PROCEDURES

Collection

     Sediment should  be collected in solvent-rinsed glass jars having TFE-
lined lids.   Each jar should be filled completely to exclude air.  A minimum
sediment sample  size of 100 g is recommended for each test.

Processing

     Sediment samples should be stored frozen at -200 c within 8 h of collec-
tion.  Holding time  should not exceed 6 mo.
                                  33

-------
                                             Laboratory Sediment Bioassays
                                              Anaphase Aberration Bioassay
                                                                 May 1986
LABORATORY  PROCEDURES
     As  previously mentioned, this technique depends on a chemical extraction
procedure.   Extraction  procedures recommended  are those presently used
by the  National  Marine Fisheries Service, National Oceanic  and Atmospheric
Administration  in Puget Sound (MacLeod  et  al. 1984).

     The following procedure is synthesized from Kocan et al.  (1982), Chapman
et al. (1982), Kocan and Powell (1985), and  Kocan et al. (1985).  Accordingly,
this procedure  supersedes previous published methodologies.

Cell  Cultures

     Although  any cell  type can  be used in this  test,  the  rainbow trout
gonad cell  (RT6-2) is recommended because  it  has been used extensively
for  this  purpose, is readily obtainable, and  is easy to cultivate.  It
has numerous mi tot ic  figures when growing exponentially and the  rmtotic
cells are  large in   comparison to those of other species,  thereby making
it easy  to count the  damaged anaphase cells.  This cell type  is  sensitive
to a wide  range of  organic chemical  compounds (Kocan et al.  1982) as well
as complex mixtures of chemicals that may  occur in marine sediments (Kocan
and  Powell  1985;  Kocan et  al. 1985).   Other cells used  for this purpose
are the  bluegil1-sunfish  line (BF-2), human  foreskin fibroblasts, newt
cells,  and  several   plant  cells (in  vivo).  Generally these cells can be
obtained from any state or federal fish disease diagnostic laboratory,
from  investigators  at both university  and federal laboratories, or from
the American Type Culture Collection (Rockville, MD).

Sediment Extraction

     Test materials  must  consist  of  substances  that  are  compatible with
the growth of cells in culture.  Generally this  can be accomplished  using
organic  extracts of environmental material such as marine sediments.

     Extractions  should be made  using  pesticide-grade reagents that have
been tested for toxicity to the cells  prior  to  their use for extraction.
This  initial  toxicity testing can be done by evaporating  a volume of the
reagent  equivalent to that which would  be  used  for the actual extraction,
and  adding  this to the cultures in varying amounts dissolved  in  the solvent
(e.g., dimethyl sulfoxide, DMSO)  to  be  used  for cell  exposure.   If the
maximum anticipated amount of the solvent blank to be used in the final
test does not significantly affect the cell  cultures, the extraction can
proceed without concern  about the possible toxic effect of the solvents.
Once  the extracts have  been made, gravimetric determinations of their  absolute
organic content must be made so that comparable organic concentrations from
each site can be used in cell cultures  exposed  to extracts  from  different
locations.   This  is accomplished by first weighing the tube  to be used
for extract storage  to  the nearest 0.001  g,  adding the extract  in the  solvent,
evaporating the solvent with nitrogen,  and reweighing the tubes.  The difference

                                  34

-------
                                             Laboratory Sediment Bioassays
                                              Anaphase Aberration Bioassay
                                                                 May  1986


between  the  original  tube weight  and  the final  tube weight is the weight
of the sample extract.   The sample extract is then  redissolved in spectral-
grade  DMSO  so  that each  sample contains the same weight/volume ratio of
extract (e.g., mg/mL).  In  this  way  it is possible  to  expose each set of
cultures  to  exactly the  same  amount of organic extract,  thereby making
comparisons  possible on an organic content basis.  If relative toxicities
from  site to  site are needed, one can extrapolate back  to the original
organic content of each sample  and  compare sites based  on  total amounts
of organic  compounds  present. The following detailed procedures for the
preparation  of sediment  extracts  follow those of  Chapman  et al.  (1982)
and MacLeod  et al. (1984).

     Sediment samples  are frozen and  stored until  just prior to extraction.
Each sample  is then thawed  and rehomogenized by thorough stirring.  An
aliquot  of  approximately 20 g wet weight is transferred to a clean, tared
beaker,  dried  to constant weight  (80° C), desiccated,  and reweighed to
determine the percent water.  A second aliquot (approximately 150 g wet
weight)  is  transferred to a tared, solvent-cleaned, 315-mL stainless steel
centrifuge bottle with a  TFE-lined screw-cap, and weighed.  The  sample
is then serially  extracted with  pesticide-grade solvents.

     Methanol (50  mL)  is added to each  centrifuge bottle.  The bottle  is
tightly capped, shaken vigorously for  2 mm, and centrifuged at 2,000 rpm
for 5  min.   The  clean solvent is  decanted into a 1-L separatory funnel.
The procedure  is  performed twice more  and the methanol extracts are combined
in the separatory funnel, which  is then closed and  covered with aluminum foil.

     One hundred  milliliters of  a dichloromethane/methanol (2:1 v/v) solution
is added to  the centrifuge bottle, the cap is closed  tightly, and the bottle
is  shaken vigorously  for 2  min to ensure complete mixing.  The bottle  is
then placed  in a  shaker table overnight  (approximately  18 h), following
which  the  sediment is settled by  centrifugation  at  2,000 rpm for 5 min
and the solvent  decanted  into  the  separatory funnel  with the methanol.
A  second  100-mL  aliquot  of  the dichloromethane/methanol  (2:1) solution
is added, the bottle is shaken vigorously, and the  bottle is then  placed
on the shaker table for 6 h.  The sediments are again settled with centrifu-
gation and the solvents are decanted.

     The remaining sample is shaken vigorously for 2 min with approximately
30 mL  of dichloromethane.  It is  then centrifuged and  the solvent is decanted
into  the  separatory funnel.   Another 100 mL of dichloromethane is added
to the bottle, the cap  is  secured,  the bottle  is shaken vigorously,  and
the bottle  is  placed  on  the  shaker  table overnight.   The sediments are
again settled with centrifugation  and the solvent is decanted  into the
separatory  funnel.  A final 30-mL rinse of dichloromethane  is added, the
bottle is shaken  vigorously, centrifuged,  and  decanted.   The sediment  is
then discarded.
                                  35

-------
                                             Laboratory Sediment  Bioassays
                                              Anaphase Aberration Bioassay
                                                                  May  1986


     Approximately 500 ml of clean,  distilled water is added to the combined
solvents in  the separatory  funnel.   The  funnel  is carefully swirled and
inverted  (with  frequent venting)  for  2 min.  The liquid phases are allowed
to separate  and the dichloromethane  (lower)  layer is drained  into  a 500-mL
separatory  funnel.   The aqueous  layer  is re-extracted twice with  20 mL
of dichloromethane and the  remainder  is  discarded.   The  dichloromethane
fractions are combined  in  the  50-mL funnel,  transferred,  with rinsing,
back to the  1-L funnel, and re-extracted with  another 500  ml  of distilled
water.   The  dichloromethane is drained  into  the 500-mL funnel  and the aqueous
layer is extracted once  more with 20  mL  of dichloromethane.  The latter
solvent is added to the 500-mL funnel  and  the aqueous  layer is  discarded.

     The dichloromethane  is drained  from the 500-mL separatory  funnel through
approximately  20 g of combusted and washed  anhydrous sodium  sulfate that
is held  in  a  30-mL  glass  conical  centrifuge  tube  with  the tip cut  off.
The effluent  from this mini-column is discharged  into a 500-mL  Kuderna-Danish
flask  with  a  15-mL  receiver.   When  empty, the  500-mL separatory funnel
is rinsed with 20 mL of dichloromethane, which is drained through the sodium
sulfate  column  into the flask.  The column  is  washed a final  time with
10 mL of dichloromethane and drained into  the flask.

     Boiling chips are added to the  Kuderna-Danish flask and a  3-ball  Snyder
column  is placed on top.   The solvent volume is reduced to about 5  mL using
a  hot  water bath.  When cooled,  the sides of  the  flask are  rinsed  into
the receiver with dichloromethane.   The receiver is removed and the contents
are  quantitatively  transferred to a tared  conical centrifuge  tube with
a ground glass stopper.  The sample is taken  almost to dryness using the
hot  water bath,  and stored  wrapped  in  aluminum foil in a desiccator with
the stopper open slightly  until a constant  weight is achieved upon reweighing
the tube. This weight is the amount of extractable organic material.

     After  weighing, the tube is closed and wrapped fully in  aluminum foil,
ready for anaphase aberration  testing.   Extracts are  treated with  1 mL
of spectrophotometric-grade DMSO for 24 h  with frequent stirring on a vortex
mixer.  The DMSO is  then removed to a glass  vial and used as "stock"  solution.
Because  all extracted material  is not dissolved  in the DMSO during testing,
the centrifuge tubes are dried and reweighed to determine  the  exact amount
used in  testing  (fraction  soluble).   Both stock and extract solutions are
stored  in the  dark under  nitrogen until applied  to the cell cultures.

Culture Conditions

     RTG-2  cells  as  well as most fish cell lines grow in a variety of com-
mercially available culture media.  The  Leibovitz L-15  medium was  found
to be  most  consistent  in terms of ease of preparation, use without special
buffers or carbon dioxide incubators, and  long-term storage capability.
This medium can  be obtained from any scientific supply house that carries
cell/tissue  culture materials.   The medium comes as a dry powder that is
added  to distilled  water,  autoclaved or  filter-steril ized,  and  stored  in

                                  36

-------
                                             Laboratory  Sediment Bioassays
                                              Anaphase Aberration Bioassay
                                                                 May 1986


a refrigerator at 4° C until  it  is used.  Full instructions are available
with the medium when it is purchased.   Generally,  heat-deactivated  fetal
calf serum  is  added to any  culture  medium at 10 percent  concentration to
ensure  that  the proper growth factors are  present.   This serum can  also
be purchased  in  100-mL to 1-L  lots already sterilized from the scientific
suppliers.   Before the cells are  placed in the medium, pH  should be adjusted
to 7.1-7.3  using either  sodium bicarbonate or HC1.   Some  laboratories use
HEPES buffer (either acid  or base in a commercially available product)
in place of the  bicarbonate  or  HC1.  This culture system differs from that
originally described by Kocan et al.  (1982) only because it has been modified
to simplify  the laboratory procedures.

Bioassay Procedure

     Cells  are grown  and tested at 180 c  on standard, clean  microscope
slides  or on 1x5  cm coverslips in Leighton tubes,  depending on the  amount
of test  material available.  The  cells are placed into  the culture system
1 day prior  to  the actual exposure to ensure  that  they have had a  chance
to attach to  the glass substrate  and begin growing.  On the following day
(18-24  h later),  the culture  medium is removed and  the  test material  is
added.   This  should consist  of normal  L-15 medium dissolved  in DMSO, to
which the organic extract has been  added.   The DMSO should be  from a pretested
lot to  ensure  that it is nontoxic to the  system and should not exceed 0.5 per-
cent (v/v)  of  the culture medium  (e.g., 5  uL/mL).   This can be  reduced
to 0.1  percent DMSO if toxicity  is  a problem or if minimal  DMSO is required
for conservation  of extract.  Exposure time  should be 48  h from the time
of addition of the treated  medium until  fixation.  Damaged cells can be
observed for longer periods even  after the toxic substance has been removed,
but the maximum response does not increase beyond 48 h.

     An  initial  screening test must be conducted to determine the actual
extract dilutions to be used for  this bioassay.  Ideally,  dilutions  tested
for  anaphase  aberrations comprise both the  highest concentration of extract
(mg/L)  that  permits continued cell proliferation (i.e.,  nontoxic)  and a
second concentration one dilution lower.  This method  ensures that  a sufficient
number  of mitotic figures  are  present to  score  for chromosome damage.
Based  on previous experience in Puget Sound,  the  following six extract
dilutions should  be  prepared:  50, 25,  10,  5,  2 and 1  ug/mL.  Cells are
first  exposed to these concentrations for  each  sediment extract tested,
and then the concentrations that inhibit  mitosis are determined.   Although
identical  extract dilutions may not be used to test each  sediment sample,
all results  are normalized to organic content  which has  been  previously
determined.

     To  determine mitotic effects and  anaphase  aberrations,  the slides
or coverslips  containing  the cells are removed from the culture  medium
and  fixed  in methanol:acetic acid (3:1).   The  methanol  is absolute and
the acetic  acid is glacial (undiluted).   Following 15-60 min in the fixative
(no adverse  effects occur  if they are  left longer) the slides  are air dried,

                                   37

-------
                                             Laboratory Sediment Bioassays
                                              Anaphase Aberration Bioassay
                                                                 May 1986


then placed  into 3 percent  Gurr's R66 Geimsa  stain for 15-30 min.  This
stain is made up in Sorensen's buffer (pH  6.8).  Optimum staining  is determined
empirically  by examining the slides with a microscope at various intervals
after placing  them  in  the  stain.   Problems  usually occur because  of too
little  staining  time rather than  excessive staining.  If excessive staining
does occur, the cells can be destained  in Sorensen's buffer.   The staining
system  selectively stains the  condensed chromosomes undergoing mitosis
and has very  little effect on  the cytoplasm of the cell, which  allows good
resolution of the small  fragments of chromosome that are not associated
with the main  chromosome bundles.

     Once the  cells have been  stained,  they are mounted on microscope  slides
(if on coverslips) or  are covered with coverslips (if  grown on  slides)
to  facilitate microscopic examination and scoring.  Slide identification
labels are covered with  a piece of tape  to  prevent observer bias  while
scoring.  Three  replicate slides are made of each exposure  concentration
with two concentrations  for  each sediment  extract.  Each slide  is then
examined at 500X  to l.OOOX until  a minimum  of 100 anaphase cells is observed
and scored.   In  this  way, there will be  three  replicates per dose with
100 anaphase  cells  per  replicate.

     The numbers  and percents of normal and abnormal anaphases  are recorded.
Cells are scored  as abnormal if they contain any of the previously described
chromosomal  lesions  reported  for this test  (Nichols et al. 1977;  Kocan
et al.  1982;  Chapman  et  al.  1982a; Kocan and  Powell 1985;  Kocan  et al.
1985).

Controls

     Controls  consist of  1) untreated cultures used as negative controls
to ensure that the  culture conditions (e.g.,  medium, serum) are  not toxic
to the cells,  2)  a  solvent blank to ensure  that the residue from the solvents
used during the extraction  procedure is  not  cytotoxic or genotoxic  (generally
done prior  to actual  testing  and includes  the DMSO that will be used  in
the final solution), and 3) a positive control consisting of cultures exposed
to  several  concentrations of  a known genotoxic agent to indicate that the
test system is functioning properly and does  indeed respond  to  genotoxic
substances.   One possible positive control is a 0.25-ug/mL  concentration
of benzo(a)pyrene.  This  level  of exposure  should result  in an  anaphase
aberration frequency of 50-65 percent (Chapman et al. 1982a).

DATA REPORTING REQUIREMENTS

     The  following data  should  be  reported  by all laboratories performing
this bioassay:

     t    Initial  screening data for the determination of extract
          concentrations
                                    38

-------
                                         Laboratory Sediment Bioassays
                                          Anaphase Aberration Bioassay
                                                              May 1986
•    Individual  replicate  and mean  and  standard deviation data
     for numbers  of normal  and  abnormal  anaphases observed

•    Types of anaphase aberrations  observed

•    Frequency of anaphase  aberrations  observed with the positive
     control

•    Any problems that may  have influenced data quality.
                               39

-------
                                             Laboratory  Sediment Bioassays
                                       Microtox Bioassay - Organic Extract
                                                                 May 1986
               MICROTOX  SEDIMENT BIOASSAY (ORGANIC  EXTRACT)
USE AND LIMITATIONS

     The Microtox bioassay is a rapid, sensitive method  of toxicity testing
based on light emission by the luminescent bacterium (Photobacterium  phos-
phor eum)  in  the presence and  absence of aqueous  toxicants.  The emitted
light is a  product of the bacterial electron  transport system  and thus
directly reflects the metabolic state of the cell.  Accordingly, decreased
luminescence  following exposure to chemical contaminants  provides a  quanti-
tative  measure  of toxicity.   The assay was developed for use  in freshwater
habitats to assess the toxicity of waterborne  pollutants  (Bulich et al.
1981)  and  has  been adapted  for use in the marine environment to  assess
toxicity of organic sediment extracts (Schiewe et al. 1985).

     This  sediment bioassay  is used to characterize the toxicity of marine
sediments.  It may be used alone as a screening tool in broad-scale  sediment
surveys,  in  combination with  sediment chemistry and  in situ biological
indices, and  in  laboratory experiments addressing a variety of  sediment
and water quality manipulations.

     The Microtox bioassay can be used with any type of sediment, regardless
of interstitial salinity or grain size characteristics.  However,  the following
caveats apply:

     •    The bioassay  depends  on a  chemical  extraction procedure
          that is specific for  neutral,  nonionic organic compounds.
          Other  classes of contaminants  such as  metals and highly
          acidic  and basic organic compounds are not  efficiently ex-
          tracted.  Thus, characterization of  sediment toxicity is
          directed towards neutral compounds  such as aromatic  and
          chlorinated hydrocarbons.

     •    Naturally occurring toxic substances may be present in  and
          extracted from marine  sediments.  Hence,  relatively high
          toxicity occasionally may be  noted  in areas generally regarded
          as  free from chemical contamination.
                                     40

-------
                                              Laboratory Sediment  Bioassays
                                        Microtox Bioassay -  Organic  Extract
                                                                  May  1986
FIELD PROCEDURES

Collection

     Sediment  should be collected  in solvent-rinsed glass jars  having  TFE-
lined lids.   Each jar should  be  filled completely to exclude air.   A minimum
sediment sample size of  500 g is recommended for each test.

Processing

     Sediment samples should  be  stored frozen at -200 c within 8  h of collec-
tion.  Holding time should not exceed 6 mo.

LABORATORY PROCEDURES

Facilities and Equipment

     The bioassay is performed using the Beckman Model 2055 Microtox Toxicity
Analyzer System,  a temperature-regulated photometer equipped  with a photo-
multiplier.   Freeze-dried bacteria,  reconstitution solution (i.e., organic-free
distilled water), diluent, and other necessary  materials  can be purchased
from  Microbics  Corporation, Carlsbad, California.  The test procedure is
conceptually quite straightforward.   However,  the methodology requires
careful attention to detail and  requires 1-2 wk to become technically  profi-
cient.

Sediment Extraction

     As previously mentioned, this technique depends on a chemical extraction
procedure.  Extraction  procedures  recommended  are those  presently  used
by the National  Marine Fisheries Service, National Oceanic and  Atmospheric
Administration in Puget  Sound (MacLeod et  al. 1984).

     •    Thaw sediment, decant and discard excess water.  Homogenize
          sediment with  stainless steel  spoon.  Discard large  pebbles,
          shells, seaweed, wood, crabs,  etc.

     t    Weigh  out 10  +0.5  g sediment  to  the  nearest 0.01 g and place
          in a dichloromethane-rinsed centrifuge bottle

     •    Set  aside approximately  10 g  of  the  homogenized sediment
          for the determination  of dry weight

     t    Centrifuge sediment sample for 5  min  at 1,000 x g and discard
          water

     •    Add  100 mL spectral-grade dichloromethane and 50 g sodium
          sulfate to sediment sample

                                   41

-------
                                    Laboratory  Sediment Bioassays
                              Microtox Bioassay  - Organic Extract
                                                        May 1986
Manually shake  bottle until contents are loose and  free-flowing
and then roll 16  h  (overnight) on a tumbler
Centrifuge sediment 5 min at 1,000 x g and  save  the dichloro-
methane extract
Add another  100 mL dichloromethane to sediment and tumble
for 6 h (during day)
Centrifuge for  5  min at 1,000 x g and save  the dichlorcmethane
extract
Repeat  extraction a third  time with 100  mL dichloromethane
for 16 h (overnight) and collect extract by centrifugation
Combine the  three 100-mL  portions  of  the dichloromethane
extract
Add  three or  four boiling  chips  to flask containing the
dichloromethane extract and attach to Synder column
Concentrate  extract to  10-15 mL in a  600 c water bath and
then transfer it  to a concentrator tube
Wash  down the  flask two  times with 3-4  mL dichloromethane
and add washings  to the tube
Add a  boiling  chip to  the  tube and,  using  a tube  heater,
concentrate the extract to greater than 0.9 but  less than
1.0 mL
Adjust  volume to  1.0 mL with dichloromethane  (at  this  point
0.1 mL may be removed for 6C/MS analyses)
Add 3  mL of hexane to  the  remaining  0.9 mL  of  extract and
concentrate to  2  mL
Remove  100  uL of extract and  add to a concentrator tube
containing 3 mL ethanol
Place tube in heater block and concentrate  to 2  mL
Adjust  volume to  3.0 mL with ethanol (the standard  sediment
extract used in Microtox  testing)
Extraction blanks are  prepared using an identical  procedure
but without the sediment.
                         42

-------
                                              Laboratory Sediment Bioassays
                                        Microtox  Bioassay - Organic Extract
                                                                  May 1986
Bioassay Procedure

     The organic  extract approach uses the basic Microtox method described
in the Microtox  Operating Manual (Beckman Instruments  1982)  and by Bulich
et al. (1981).

     •    Reconstitute lyophilized bacteria with  1-mL double-distilled,
          charcoal-filtered water  in  a Microtox cuvette  and place
          in 40 c holding well.  Bacterial suspensions are used within
          8 h of reconstitution.

     •    Prepare three or more widely spaced  primary dilutions (e.g.,
          5.0, 0.5,  0.05 percent extract  [v/v])  in  double-distilled,
          charcoal-filtered water.

     •    Adjust concentration of  each primary dilution to 2 percent
          NaCl by adding  0.1 ml 22  percent NaCl per mL  of diluted
          extract.

     •    Use these  diluted extracts in a range-finding assay to determine
          an appropriate primary dilution  for  the definitive  assay
          described  below.  The primary dilution should cause a 65-90 per-
          cent decrease in  bioluminescence  in  15  min.  Methods  for
          the range-finding assay are the same  as those for the definitive
          assay  except usually  only three concentrations are tested
          without  replication.

     •    For the definitive assay,  two-fold serial dilutions (i.e.,
          5.0, 2.5,  1.25, and 0.625 percent) of  extract are  prepared
          in 2 percent  NaCl.  A 2 percent NaCl  blank  is also  prepared
          for testing  to measure spontaneous decay of  light production
          which  occurs naturally independent of treatment.

     0    In each of  10  test cuvettes, a 10 uL aliquot of bacterial
          suspension is added to 500 uL  of diluent  and incubated  for
          15 min  in the  incubation wells.  This assures temperature
          equilibration and  stability of bioluminescence.

     •    After 15  min,  initial levels of light emission are  measured
          in each  of the 10  test cuvettes.

     •    At 30-sec intervals,  500 ml al iquots of each concentration
          of extract  are added  to two  of the cuvettes (i.e., two replicates
          each  of the  four extract dilutions  and the  saline blank).
          Timing is critical  because  bioluminescence  gradually decreases
          over time.
                                   43

-------
                                             Laboratory Sediment Bioassays
                                       Microtox Bioassay - Organic Extract
                                                                 May 1986
          Exactly 5 min after  addition of the sediment extract, light
          emission  is  measured at 30-sec  intervals and  in  the same
          sequence used for  extract additions in the  preceding step.
          Light emission is measured  again at 15 min and 30  min.

          Immediately after testing  each sediment extract,  an  ethanol-
          only control is  assayed using the  same primary dilution
          sequence used in the sediment extract test.  The  ethanol-only
          data are used to adjust  the  sediment extract data  for  the
          contribution of the solvent vehicle.

          Estimates of the  15-min EC50s (i.e.,  the concentration of
          extract causing a 50 percent  reduction  in  bioluminescence)
          are obtained using  linear regression analyses (see  Microtox
          Operating Manual).  Briefly, the percent inhibition  of light
          emitted at each test concentration and time point is  converted
          to  a gamma  value  which is defined  as  the ratio of  light
          lost to light remaining.   Gamma values are normal i zed for
          natural decline in light production measured over  time as
          described by Bulich  et al.  (1981) and further  adjusted for
          the contribution of the  ethanol  vehicle.  The  natural  log
          of  gamma is  regressed  on the natural log of extract concentration
          and the  EC50 is calculated  from  the regression equation.
          A  statistical procedure based on Fieller's theorem (Finney
          1964) is used to calculate  a 95-percent confidence interval.
Controls
     •    Ethanol, sodium  lauryl  sulfate, or other suitable  compound
          should be used to  assess daily bioassay performance  and
          to determine differences  in  response among lots of bacteria.

     •    Bioassay repeatability  is evaluated  by duplicate  testing
          of 10  percent of the sediment  extracts.

DATA REPORTING REQUIREMENTS

     The following  data should be  reported by all laboratories performing
this bioassay:

     •    Range-finding assay results

     t    Raw light emission data for  each test series

     •    15-min EC50 data  and 95-percent confidence  intervals  for
          each test series and for  controls

     •    Any problems that may have influenced data quality.

                                  44

-------
                                             Laboratory Sediment Bioassays
                                         Microtox Bioassay - Saline Extract
                                                                  May  1986
                    MICROTOX  BIOASSAY  (SALINE EXTRACT)
USE AND LIMITATIONS
     The use of  saline extracts  of sediment  for  the Microtox bioassay has
been  described by Williams et al. (in press).   This technique is not presently
recommended  for widespread use in Puget Sound as  it requires further research
to evaluate  test  variables and effects of sediment storage  and extraction
techniques  on  the precision of the  results.   However, the techniques are
well  developed  and  are  provided herein for ease of use  in  possible future
research efforts.

     The organic  and saline extract approaches  both use the basic Microtox
method described  in  the  Microtox Operating Manual (Beckman Instruments
1982) and by Bulich et  al. (1981).  The major difference  is in  the preparation
of test samples.   Each  procedure  is  specific  with regard  to the  classes
of contaminants that  are  tested for toxicity and, in general, the results
for each approach  can be viewed as complementary.

     A limitation of the saline  extract Microtox  bioassay  procedure is
that only the water-soluble fraction of sediment-adsorbed  trace metals
and organic  pollutants  are extracted from the sediments.  Thus, contaminants
with extremely  low water  solubilities  (PCBs) will tend  to be partitioned
almost  exclusively  onto  sediment  particles  and are unlikely to occur in
high concentration in the saline extract used in  toxicity testing.   Another
limitation  is  the need  to  establish a  correction factor  for changes in
bacterial luminescence  caused by variation among  samples  in  sediment  pore-
water  salinity.   Although Williams et al.  (in press) showed that salinity-
induced changes in luminescence were negligible for sediments taken  from
Commencement Bay, other  estuarine  sediments may have a greater range of
pore-water salinities  and may  require a salinity correction factor.  Finally,
the assay requires further research to demonstrate  precision of the extraction
procedure and stability of sediment  toxicity during periods  of  prolonged
sediment storage (i.e., 2 wk).

FIELD PROCEDURES

Collection

     Sediment  should be collected  in solvent-rinsed glass jars having TFE-
lined lids.   Each jar should  be filled completely to exclude air.  A minimum
sediment sample size  of 200 g is recommended  for  each test.
                                  45

-------
                                             Laboratory Sediment Bioassays
                                        Microtox Bioassay  -  Saline Extract
                                                                 May 1986
Processing
     Sediment samples  should be stored at 40 c in the dark.   Holding time
should not  exceed 14 days.

LABORATORY  PROCEDURES

     With slight modification, sample  processing follows procedures described
by Williams et al.  (in  press).   Precleaned  glass test tubes  are  filled
with  a minimum  of 60 g of sediment,  sparged with high purity  nitrogen gas,
tightly  sealed with TFE-lined caps, stored  in the dark at 40 C,  and assayed
within 14 days following collection.

Preparation of Sediment Extract

     •    Remove 30 g of sediment from  each test tube with a stainless
          steel  spatula,  place  in 30-mL glass  containers equipped
          with a fritted glass cap, and add 10.0 mL of Mirotox diluent
          (2.0 percent  NaCl  w/v in double-distilled, organic-free
          water).

     •    Wash the  sediment-diluent slurry  for 24 h in  the  dark at
          40 c by gentle agitation (100  rpm) on a rotary shaker table.

     •    Transfer  the sediment slurry to 30-mL Corex tubes and centrifuge
          for 15 min at 9,000 rpm (9,770 x  g)  in a refrigerated (40 C)
          centrifuge.

     •    Draw the  supernatant  off by pipette, place  it  in  a clean
          test tube, cool on  ice  and  use immediately in preparation
          of serial  dilutions for the  Microtox bioassay.

Bioassay  Procedure

     •    Rehydrate a vial  of freeze-dried  bacteria with 1.0 mL of
          reconstitution solution,  cover  with  parafilm, store in  a
          40 C well on  the  Microtox analyzer,  and use within 5 h of
          rehydration.

     •    Prepare  100,  50,  25,  12.5 and 0 percent serial dilutions
          of the sediment supernatant in Microtox diluent.  The 0 percent
          dilution  is a  reagent blank  needed to measure spontaneous
          decay  in  bacterial luminescence independent of any treatment.

     t    In each  of 10  test cuvettes, add  10  uL of the  rehydrated
          bacterial  suspension to 350  uL of diluent  and incubate for
          15 min  in one of  the 150  c wells  on  the analyzer.  This
          assures  temperature equilibration  of the  bacterial suspension
          and stability of luminescence.

                                  46

-------
                                             Laboratory Sediment Bioassays
                                        Microtox Bioassay  -  Saline Extract
                                                                 May 1986
     •    After  15 min, measure  initial luminescence in  each  of the
         10 test cuvettes.

     •    At 30-sec intervals,  add  500 uL aliquots of each  supernatant
         dilution to two of the cuvettes (e.g., two  replicates each
         of the four test  dilutions and the  saline blank).   Timing
         is critical because toxicant-induced decrease in luminescence
         begins as soon as the  sediment supernatant is added  to the
         bacterial suspension.

     •    Exactly 15 min after addition of the sediment supernatants,
         measure  luminescence at  30-sec  intervals  and in the same
         sequence used for supernatant additions in  the  preceding
         step.

     t    Calculate percent decrease in luminescence relative  to the
         reagent blank using the  formula:

                   Percent Decrease =  [(RI0-It)/(RIo)] x 100

     where:

         I0 = initial luminescence
         It = luminescence at the end  of 15 min
         R  = blank ratio.

     The  blank ratio is calculated by:

                                 R =  Bt/Bo

     where:

         B0 = initial luminescence  of the reagent blank
         Bt = luminescence of the reagent blank after 15 min.

Controls

     •   Clean  reference sediments  used  as negative controls.

     t   Construction  of  a calibration curve to determine salinity
          induced changes in bacterial  luminescence.

     •   Use  of sodium arsenate  as a  reference toxicant to  assess
         day-to-day performance  of the bioassay and  to  determine
         differences in toxic response among  lot numbers of bacteria.

     •   Verification of a dose-response relationship between bacterial
         luminescence and sediment  extract concentration.

                                  47

-------
                                              Laboratory Sediment  Bioassays
                                         Microtox Bioassay  -  Saline Extract
                                                                  May 1986
DATA REPORTING REQUIREMENTS

     The following data should  be reported by all  laboratories  performing
this bioassay:

     •    Percent decrease  in  luminescence for each  concentration
          of supernatant  (e.g., saline sediment extract)  tested.

     •    Determination  of a significant dose-response relationship
          by least-squares regression of percent decrease in luminescence
          on the logarithm  of sample dilution.

     •    Determination of  EC50 values and 95-percent confidence  limits
          for the reference toxicant.
                                     48

-------
                                             Laboratory Sediment  Bioassays
                                                Other Promising Techniques
                                                                 May 1986
                        OTHER  PROMISING TECHNIQUES


     This section  describes  promising  sediment bioassay techniques that
may be generally  applicable in Puget Sound  following further detailed  testing
and validation of the methods.   The techniques are listed in terms of relative
priority:

     1.    The echinoderm sperm  bioassay has been  used in sediment
          bioassays  using elutriates but may be affected by the physical
          presence of  particles, masking any chemical toxicity (Ross
          et al. 1984; Malins  et  al.  1985; Dinnel  and Stober  1985).
          If this test can be successfully adapted to sediment bioassay
          testing,  it should prove useful.  Echinoderms such  as sea
          urchins  and sand dollars  may be usable  interchangeably,
          allowing testing  to occur  over most  of the year.   Draft
          protocols have  been developed  by the U.S. EPA (Narragansett
          Laboratories) for conducting effluent toxicity tests with
          the sea  urchin  Arbacia punctulata  (Nacci et al. in review).
          These  protocols may be appl icable  in large  measure to  Puget
          Sound sediment  testing. In addition to the echinoderm sperm
          bioassay, the echinoderm embryo  bioassay described by  Dinnel
          and Stober  (1985) may also  be  applicable to sediment testing
          (Ross  et  al. 1984),  allowing  a wider range of life-cycle
          testing for this group of organisms.

     2.    Extension of  the  bivalve larvae sediment bioassay technique
          described earlier to bivalve species other  than the Pacific
          oyster  and blue mussel should be possible.  ASTM  (1985)
          has standardized procedures  for  only four  species:  Pacific
          oyster (Crassostrea gigas), blue  mussel (Mytilus edulis).
          Eastern oyster  (Crassostrea virginica), and quanog (Mercenaria
          mercenaria).  APHA  (1985)  only discusses larvae tests with
          oysters.  However, Puget Sound  species such  as the Olympia
          oyster (Ostrea lurida), Macoma  balthica, and geoduck (Panopea
          generosa) may be useful alternative species  for use  in the
          bivalve larvae sediment bioassay.

     3.    Three techniques  other than  the echinoderm  sperm test that
          were listed  in Table 3  but  not recommended  for widespread
          use in Puget  Sound  are:  1) the oligochaete, Monopylephorus
          cuticulatus. respiration rate test; 2)  the surf  smelt, Hypomesus
          pretiosus pretious, partial  life-cycle test;  and,  3)  the
          copepod.  Tigriopus californicus, partial  life-cycle test.
          These  tests are  all  sensitive methods, but are not immediately
          applicable.   The oligochaete respiration test  and the copepod

                                  49

-------
                                        Laboratory Sediment Bioassays
                                           Other Promising Techniques
                                                            May 1986
     partial life-cycle test are highly  promising tests  that,
     if  further developed, may be  suitable for inclusion in future
     updates  of this document.   The surf smelt  egg  to larvae
     test may be useful if natural variability  can be  overcome.
     In  any case,  some  type of  fish egg development test (e.g.,
     herring or cod) would be extremely valuable  for future use
     in  Puget Sound, even though  it is recognized that such species
     cannot  be tested year-round.

4.   Life-cycle bioassays with nematodes as mentioned in Table 2,
     although not  used  to date  in Puget  Sound, may  provide a
     rapid   and effective means  of determining chronic toxicity
     of  sediments.

5.   The U.S. EPA (Narragansett Laboratories) is presently developing
     a solid phase  sediment toxicity  test  using the  infaunal
     amphipod Ampelisca abdita (Gentile  et al. 1985).  This species
     apparently is similar to Rhepoxynius abronius with respect
     to  toxicant sensitivity.  However, it is  found in fine sediments,
     whereas £. abronius  is  not  (Scott, K.J.,  31  October  1985,
     personal communication).   Ampelisca  spp. also occur  along
     the west coast  of  North America,  and  this  animal may be
     a useful species for testing those fine sediments that may
     have some measure of physical effect on R. abronius.

6.   The U.S. EPA (Narragansett Laboratories) is presently finalizing
     testing protocols for  complex  effluents  in  marine waters.
     Four such tests have been developed: 1) 7-day survival/growth/
     reproduction with  the  mysid,  Mysidopsis  bahia;  2) larval
     fish  growth/survival  with  the sheepshead minnow,  Cyprinodon
     variegatus;  3) sea  urchin  (Arbacia punculata) sperm  eel 1
     tests;  and 4)  sexual reproduction tests with the alga Champia
     parvula.  The sea urchin sperm cell test  is  already  noted
     in  this document as a promising technique.  The other  three
     tests may be useful methods  to  test the toxicity of sediment
     elutriates.

7.   The Federal Register (U.S. EPA  1985)  recently  introduced
     formal  test guidelines for several  types  of bioassays  under
     the Toxic Substances Control Act.   Two of these tests are
     not listed  above:  algae and penaeid  shrimp  acute  tests.
     These  may be useful methods to test the toxicity of sediment
     elutriates.
                              50

-------
                                REFERENCES


APHA.  1985.   Standard methods for the examination of water  and wastewater.
16th edition.   1268 pp.

ASTM.   1985.   Annual book of ASTM standards.  Uater and  Environmental Tech-
nology.  Volume 11.04.   American Society for Testing and  Materials,  Phila-
delphia, PA.   pp.  256-272.

Battelle.   1986.   Reconnaissance level  assessment  of selected  sediments
from Puget Sound.   Report prepared for U.S. Environmental Protection Agency,
Seattle, WA.   229 pp.

Beckman Instruments, Inc.  1982.  Microtox system operating manual.  Carlsbad,
CA.

Bulich,  A.A., M.W.  Greene, and D.L.  Isenberg.  1981.   Reliability of the
bacterial luminescence  assay for determination  of the toxicity of  pure
compounds and complex  effluent,  pp.  338-347.   In:   Aquatic Toxicology
and Hazard Assessment:   Proceedings of the Fourth  Annual Symposium.   ASTM
STP 737.  D.R. Branson and K. L. Dickson (eds). American  Society  for Testing
and Materials, Philadelphia, PA.

Chapman, P.M.   (In press.)  Oligochaete respiration as a  measure  of toxicity
in Puget Sound, Washington.  In:  Proceedings  of  the Third  International
Symposium on Aquatic Oligochaete Biology, Hamburg, Germany,  Sept. 30-Oct. 5,
1985 (Mitt.  Hamb.  Zool.  Mus. Inst.).

Chapman, P.M., and  R.  Fink.   1983.  Additional marine sediment toxicity
tests  in connection with toxicant  pretreatment  planning studies,  Metro
Seattle.  Unpublished report prepared by E.V.S. Consultants for  the Munici-
pality of Metropolitan Seattle.

Chapman, P.M., and  J.D.  Morgan.  1983.   Sediment  bioassays with oyster
larvae.  Bull. Environ.  Contam. Toxicol. 31:438-444.

Chapman, P.M., and  C.T.  Barlow.  1984.   Sediment bioassays in various BC
coastal areas.  Unpublished  report prepared by  E.V.S. Consultants for the
Environmental  Protection Service.

Chapman, P.M., and  R.  Fink.  1984.  Life-cycle studies  with  the polychaete
Capitella capitata exposed to contaminated whole sediments  and  elutriates.
Bull.  Environ. Contam. Toxicol. 33:451-459.

Chapman,  P.M.,  G.A. Vigers, M.A. Fanrell, R.N. Dexter, E.A. Quinlan,  R.M. Kocan,
and  M.L. Landolt.   1982a.   Survey of biological effects of toxicants upon
Puget Sound biota.  I. Broad-scale toxicity survey.  NOAA Tech.  Memo. OMPA-25.
98 pp.
                                   51

-------
Chapman,  P.M., M.A.  Parrel 1. R.M. Kocan, and  M.L.  Landolt.  1982b.  Marine
sediment toxicity  tests  in connection with toxicant  pre-treatment  planning
studies,  Metro Seattle.   Report prepared  by E.V.S. Consultants for the
Municipality of Seattle.

Chapman,  P.M., O.R.  Munday, J. Morgan, R.  Fink,  R.M. Kocan, M.L. Landolt,
and R.N. Dexter.   1983.   Survey of  biological effects  of toxicants  upon
Puget Sound  biota.   II.  Tests of reproductive  impairment.  NOAA Tech. Report
NOS 102 OMS-1.   58 pp.

Chapman,  P.M., R.N.  Dexter, J. Morgan, R.  Fink,  D. Mitchell, R.M. Kocan,
and M.L. Landolt.   1984.  Survey of  biological effects  of toxicants  upon
Puget  Sound biota.   III.   Tests in Everett Harbor,  Samish, and Bellingham
bays.  NOAA  Tech.  Memo.  NOS OMS-2.  48 pp.

Chapman,  P.M., R.N.  Dexter, R.M. Kocan, and E.R.  Long.   1985.  An overview
of biological effects  testing in Puget Sound, Washington;  methods,  results,
and implications,   pp. 344-363.  In:  Aquatic Toxicology and Hazard Assessment:
Proceedings  of the Seventh Annual Symposium ASTM STP 854.  R.D.  Cardwell,
R. Purdy, and R.C.  Bahner (eds).  American Society  for Testing and Materials,
Philadelphia, PA.

Cummins,  J.M., and C.E.  Gangmark.   (in prep.)   Monitoring changes in the
sensitivity of wild populations of the marine amphipod, Rhepoxynius  abronius.
used  in sediment  toxicity tests.   U.S. Environmental  Protection Agency,
Manchester,  WA.

Cummins,  J.M., R.R.  Bauer, R.H. Rieck, W.B. Schmidt,  and J.R.  Yearsley.
1976.  Chemical and biological survey of  Liberty Bay,  Washington.   EPA-910/
9-76-029.  U.S. Environmental Protection Agency, Washington, DC.  132 pp.

Dinnell, P.A., and Q.J.  Stober.   1985.   Methodology and analysis  of sea
urchin embryo bioassays.  Fisheries  Research Institute (FRI) Circular No.  85-3.
19 pp.

Finney, D.J.  1984.  Statistical  methods in biological  assay.  Charles
Griffin and  Company Limited,  London.

Gentile, J.H., K.J. Scott,  S. Lussier, and M.S. Redmond.   1985.  Application
of laboratory population responses  for evaluating  the  effects of dredged
material. Technical  Report D-85-8.   Prepared by the  U.S.  Environmental
Protection Agency, Environmental  Research Laboratory, Narragansett, RI
for the U.S. Army  Engineer  Waterways Experiment Station,  U.S. COE,  Vicksburg,
MS.

Hargis, W.J., M.H. Roberts,  and  D.E. Zwerner.  1984.  Effects of contaminated
sediments and sediment-exposed effluent water on an estuarine  fish:   acute
toxicity.  Mar. Environ. Res. 14:337-354.

Kocan,  R.M., M.L.  Landolt,  and  K.M. Sabo.  1982.   Anaphase aberrations:
a measure of genotoxicity  in  mutagen-treated fish cells.   Environ. Mutagenesis
4:181-189.
                                  52

-------
Kocan,  R.M., and D.B.  Powell.  1985.   Anaphase aberrations: an  in vitro
test for assessing the genotoxicity  of individual  chemicals  and  complex
mixtures,   pp.  75-86.   In:   Short-term bioassays in  the analysis of complex
environmental  mixtures.   IV.  M.D. Waters, S.S. Sandhu, J. Lewtas, L.  Claxton,
G.  Strauss,  and  S. Nesnow (eds).   Plenum Publishing Corp.,  New York, NY.

Kocan, R.M.,  K.N.  Sabo,  and M.L. Landolt.  1985.  Cytotoxicity/genotoxicity,
the application  of cell  culture  techniques  to  the measurement of marine
sediment pollution.  Aquat. Toxicol. 6:165-177.

Landolt,  M.L., and R.M.  Kocan.   1984a.   Lethal and sublethal effects of
marine sediment extracts  on  fish cells and chromosomes.  Helgol.  Meeresunters
38:125-139.

Landolt,  M.L., and R.M.  Kocan.   1984b.   Anaphase  aberrations in  cultured
fish cells as  a bioassay  of  marine sediments.  Mar. Environ.  Res. 14:497-498.

Lee,  C.R., B.L.  Flosom,  Jr., and R.M.  Engler.  1982.  Availability and
plant uptake of heavy metals  from contaminated dredged material placed
in flooded and upland  disposal  environments.  Environ. Int.  7:65-72.

Legore, R.S.,  and D.M. DesVoigne.  1973.  Absence of acute effects  on threespine
sticklebacks  (Gasterosteus  aculeatus) and coho  salmon (Oncorhynchus  kisutch)
exposed to resuspended  harbor  contaminants.  J.  Fish.  Res. Board Can. 30:124-
1242.

MacLeod,  Jr., W.D.,  D.W.  Brown, A.J. Friedman, 0. Maynes,  and R.W. Pearce.
1984a. Standard analytical procedures  of the NOAA  National Analytical Facility,
1984-1985.  Extractable  toxic organic compounds.   NOAA  Tech. Memo. NMFS
F/NWC-64.   110 pp.

MacLeod,  W.D., D.W.  Brown,  A.J.  Friedman,  0.  Maynes, and  R.W.  Pearce.
1984b.  Standard analytical  procedures of the  NOAA   (National  Oceanic and
Atmospheric Administration)  National Analytical Facility, 1984-1985:  extract-
able toxic organic compounds.   NOAA-TM-NMFS-F/NWC-64.  102 pp.

Mai ins,  D.C., S-L.  Chan,  U.  Varanasi, M.H. Schiewe,  J.E. Stein, D.W.  Brown,
M.M. Krahn, and B.B. McCain.   1985.   Bioavailability and toxicity of  sediment-
associated chemical  contaminants to marine biota.  Unpublished report prepared
by NOAA/NMFS, Seattle, WA.   30 pp.

McCain, B.B., M.S. Myers,  U.  Varanasi, D.W. Brown, L.D. Rhodes, W.D.  Gronlund,
D.G. Elliott, W.A. Palsson,  H.O. Hodgins, and D.C. Malins.  1982.   Pathology
of  two  species of  flatfish  from  urban estuaries in Puget  Sound.   EPA-600/
7-82-001.   100 pp.

Misitano, D.  1983.   Effects of  contaminated  sediments on  reproduction
of  a marine copepod.   pp.  26-28.   In:  NOAA/NMFS  Quarterly Report  Oct.-Dec.,
1983.

Nacci,  D.E., R.  Walsh, and  E. Jackim.  (In  review.)  Guidance  manual  for
conducting sperm cell  tests with  the  sea urchin, Arbacia  punctulata.  for
use in  testing complex effluents.   U.S.  Environmental Protection Agency,
Environmental Research Laboratory,  Narragansett,  RI.    16 pp.

                                   53

-------
Nichols,  W.W.,  R.C. Miller,  and C.  Bradt.   1977.   In vitro  anaphase and
metaphase  preparations in mutation testing,   pp.  225-233.   In:   Handbook
of Mutagenicity Test Procedures.   B.J. Kilbey, M.  Legator, W.W. Nichols,
and C.  Ramel  (eds).  Elsevier Scientific  Publishing Co.,  New York,  NY.

Ott, F.S.,  P.O. Plesha, R.D.  Bates,  C.  Smith, and B.B.  McCain.   (In prep.)
Developing  sediment bioassays as an ecological monitoring tool.  U.S. Dept. of
Commerce,  Natl.  Marine Fish. Serv.

Pierson,  K.B.,  B.D. Ross, C.L. Melby,  S.D.  Brewer, and  R.E.  Nakatam.
1983.   Biological testing of solid phase  and suspended phase dredge material
from Commencement Bay, Tacoma, Washington.  U.S. Army Corps  of Engineers
DACW67-82-C-0038.  59 pp.

Ross,  B.,  P.  Dinnel, and Q. Stober.   1984.  Marine toxicology,   pp. 291-370.
In:  Renton Sewage  Treatment Plant  Project:   Duwamish Head.   Q.  Stober
and K.  Chew (eds).   Fisheries Research Institute FRI-UW-8417.

Schiewe,  M.H.,  E.G.  Hawk, D.I.  Actor,  and  M.M.  Krahn.   1985.   Use of a
bacterial  bioluminescence assay to  assess toxicity of contaminated marine
sediments.   Can. J.  Fish. Aquat. Sci.  42:1244-1248.

Schink, T.D., W.E. Westley, and C.E. Woelke.  1974.  Pacific oyster embryo
bioassays  of bottom  sediments from Washington waters.  Washington State
Department of Fisheries.  24 pp.

Scott, K.J.   31 October  1985.   Personal Communication  (Contact  by  P.M.
Chapman).   U.S.  Environmental  Protection Agency, Environmental Research
Laboratory, Narragansett, RI.

Shuba, P.J., H.E. Tatum,  and J.H.   Cornall.  1978.  Biological  assessment
methods to predict the impact of open  water  disposal of dredged  material.
Dredged Material  Research Program, Tech.  Rpt.  D-78-50.  Dept. of Army,
Vicksburg, MS.   77 pp.

Swartz, R.C., W.A. DeBen, and F.A. Cole.   1979.  A bioassay for  the toxicity
of sediment to marine  macrobenthos.  J.  Water  Pollut. Control  Fed.  51:944-950.

Swartz, R.C., D.W. Schultz,  G.R.  Ditsworth, W.A.  DeBen,  and  E.A. Cole.
1981.   Sediment  toxicity, contamination,  and  benthic community  structure
near ocean disposal  sites.  Estuaries 4:258.

Swartz, R.C., W.A. DeBen, K.A. Sercu,  and J.O.  Lamberson.   1982.   Sediment
toxicity  and the distribution of amphipods  in  Commencement Bay, Washington,
U.S.A.   Mar.  Pollut.  Bull.  13:359-364.

Swartz, R.C., W.A. DeBen, O.K. Phillips,  J.O.  Lambserson, and  F.A. Cole.
1985a.  Phoxocephalid  amphipod bioassay  for marine sediment  toxicity.
pp.  284-307.  In:  Aquatic Toxicology  and  Hazard Assessment:   Proceedings
of the Seventh Annual Symposium.  ASTM STP 854.  R.D. Cardwell,  R.  Purdy,
and R.C.  Banner (eds).  American Society for  Testing and Materials, Phila-
delphia,  PA.


                                  54

-------
Swartz,  R.C.,  D.W. Schults,  G.R.  Ditsworth,  W.A. DeBen,  and  F.A.  Cole.
1985b.   Sediment toxicity,  contamination,  and macrobenthic  communities
near a large  sewage outfall,  pp. 152-175.   In:   Validation and predictability
of laboratory methods for assessing the  fate and effects  of contaminated
aquatic ecosystems.  ASTM STP 865.  T.P.  Boyle  (ed.).

Swartz,  R.C.,  G.R. Ditsworth,  D.W. Schults,  and J.O. Lamberson. 1986.
Sediment toxicity to a marine infaunal  amphipod:  cadmium and  its interaction
with sewage  sludge.  Mar. Environ. Res.   18:133-153.

Tietjen,  J.M.,  and J.J. Lee.   1984.   The  use  of  free-living nemotodes  as
a bioassay for estuarine sediments.   Mar.  Environ. Res. 11:233-252.

Tsai,  D.F.,  J.  Welch,  K.W.  Chang, J.  Shaeffer, and L.E.  Cronin.  1979.
Bioassay of  Baltimore Harbor sediments.   Estuaries 2:141-153.

U.S. Environmental Protection  Agency.   1985.   Toxic Substances Control
Act test guidelines;  final  rules.  U.S. EPA, Washington, DC.  Federal Register,
Vol. 50, No.  188, Parts 796, 797, and 798.   pp.  39252-39584.

Williams, L.G.,  P.M. Chapman, and T.C.  Ginn.  (In  press.)   A comparative
evaluation of sediment toxicity using bacterial  luminescence, oyster embryo,
and amphipod  sediment bioassays.   Mar.  Environ.  Res.

Wurster,  C.F.   1982.   Bioassays for evaluating  chemical toxicity to marine
and estuarine plankton.  Unpublished report.  NOAA/MESA  -  New York  Bight
Project Office.   13 pp.
                                   55

-------
FISH PATHOLOGY

-------
MICROBIOLOGY

-------
FINAL REPORT
TC-3991-04                        Pugef Sound Estuary Program
RECOMMENDED PROTOCOLS FOR
MICROBIOLOGICAL STUDIES
IN PUGET SOUND
Prepared by:

TETRA TECH, INC.

and

E.V.S. CONSULTANTS, INC.
Prepared for:
U.S. ENVIRONMENTAL PROTECTION AGENCY
Region  10 - Office of Puget Sound
Seattle, VVA
November, 1986
TETRA TECH, INC.
11820 Northup Way
Bellevue, WA 98005

-------
                                  CONTENTS






                                                                        Page



LIST OF TABLES                                                          iii



INTRODUCTION                                                              1




MICROBIOLOGICAL MEASUREMENTS CURRENTLY MADE IN PUGET SOUND                3



RECOMMENDATIONS FOR FUTURE STUDIES                                        5



     BACTERIAL INDICATORS                                                 5



     PRIMARY PATHOGENS                                                    5



SPECIAL SAMPLING CONSIDERATIONS                                           6



     WATER COLUMN                                                         6



     SEDIMENTS                                                            6



     TISSUE                                                               7



USES AND LIMITATIONS OF RECOMMENDED BACTERIAL INDICATORS                  8



     FECAL COLIFORM BACTERIA AND FECAL COLIFORM BACTERIA/E. COLI          8



     ENTEROCOCCI                                                         10



     CLOSTRIDIUM PERFRINGENS                                             11



LABORATORY PROCEDURES FOR RECOMMENDED BACTERIAL INDICATORS               12



QUALITY ASSURANCE/QUALITY CONTROL (QA/QC)                                14



REFERENCES                                                               15
                                  11

-------
                                  TABLES
Number

   1

   2
Contributors to  the microbiology protocols

Current bacteriological measurements in Puget  Sound
   3    Recommended  bacterial indicators for monitoring  in  Puget
        Sound

   4    Recommended  laboratory procedures for bacterial  indicators
Page

  2

  4


  9

 13
                             ACKNOWLEDGEMENTS
     This chapter was prepared  by Tetra Tech,  Inc., under the direction
of Dr. Scott Becker,  for  the U.S.  Environmental Protection Agency  (EPA)
in partial  fulfillment  of Contract  No.  68-03-1977.  Dr.  Thomas Ginn of
Tetra Tech was the Program  Manager.  Mr. John Underwood  and Dr. John Armstrong
of U.S. EPA  were the  Project Officers.  Mr. Peter  Nix  of E.V.S. Consultants,
Inc.,  was the primary author of this chapter.
                                 iii

-------
                                                       Microbiology Protocols
                                                                 Introduction
                                                                 October 1986
                               INTRODUCTION
     This chapter  presents recommendations for measuring  microorganisms
in Puget Sound.  The recommendations  are  based on the results  of  a  workshop
and on  written reviews by  representatives  from  most organizations that
fund or conduct microbiological studies in the Sound (Table 1).   The purpose
of developing  these recommendations is to encourage all  Puget  Sound investi-
gators conducting monitoring  programs, baseline  surveys,  and intensive
investigations to use  standardized methods whenever possible.  If this
goal is achieved, most data  collected in Puget  Sound  should be  directly
comparable and thereby capable of being integrated into  a  sound-wide database.
Such a database is necessary for developing and maintaining a  comprehensive
water quality  management program for  Puget Sound.

     The initial  section  of this chapter  describes  those  microorganisms
currently being measured in  Puget Sound.   In  subsequent  sections  recom-
mended microorganisms for future studies  are  identified, and special consider-
ations for sampling water,  sediment, and tissue  are discussed.   Finally,
the uses  and  limitations of  the recommended microorganisms  are discussed,
and laboratory and quality  assurance/quality  control  (QA/QC)  procedures
are recommended.

     It is recognized  that departures  from the general  recommendations
made herein may be necessary to meet  the  special requirements  of  individual
projects.   If such departures  are  made,  however,  the  funding agency or
investigator should be aware that the resulting data may not  be  comparable
with most  other data of that kind.   In some  instances,  data collected using
different methods may be compared  if the  methods  are intercalibrated  adequately.

-------
            TABLE 1.  CONTRIBUTORS TO THE MICROBIOLOGY  PROTOCOLS
Name
Organization
Carlos Abeyta
John Armstrong
Scott Becker
Norma Christopherson

Alfred Dufour
Gary Fraser

Mike Glass
Peggy Hammer
Andrew Heyward
Nancy Jensen
Lawrence Kamahele
Larry Kirchner
Jack Lilja

Steve Martin
Jack Matches
June Nakata
Peter Nix
Mike Schiewe

Jay Vasconcelos
Marleen Wekell
U.S. Food and Drug Adminstration
U.S. Environmental Protection Agency
Tetra Tech, Inc.
Washington Department of Social and
Health Services
U.S. Environmental Protection Agency
Washington Department of Social and
Health Services
Washington Department of Social and
Health Services
U.S. Environmental Protection Agency
Municipality of Metropolitan Seattle
Washington Department of Ecology
Seattle-King County Health Department
Seattle-King County Health Department
Washington Department of Social and
Health Services
U.S. Army Corps of Engineers
University of Washington
Seattle-King County Health Department
E.V.S. Consultants, Inc.
National  Oceanic and Atmospheric
Administration
U.S. Environmental Protection Agency
U.S. Food and Drug Administration

-------
                                                       Microbiology Protocols
                                                         Current Measurements
                                                                October 1986
               MICROBIOLOGICAL MEASUREMENTS CURRENTLY MADE
                              IN PUGET SOUND
     A variety of  agencies sample  for  bacteria in the water,  sediment,
or biota of  Puget Sound.  These agencies and the various  bacterial indicators
(i.e., of  pathogens) and/or  primary pathogens  they  evaluate are listed
in Table  2.  Measurements of bacterial indicators  that  are currently  used
include:

     t   Total counts of coliform bacteria (i.e.,  total  coliform
         bacteria)

     t   Counts of  fecal coliform bacteria (i.e.,  fecal  coliform
         bacteria)

     t   Counts  of Escherichia col i as a fraction of fecal coliform
         bacteria  (i.e., fecal  coliform  bacteria/E.  coli)

     t   Counts  of enterococci, a  subset of fecal streptococci (i.e.,
         enterococci)

     •   Counts of Clostridium perfringens (i.e., C. perfringens)

     0   Counts  of heterotrophic  bacteria by  standard plate count
         methods (i.e., standard plate counts).

-------
                     TABLE 2.   CURRENT BACTERIOLOGICAL MEASUREMENTS  IN  PUGET  SOUND*
Agency
               Routine Monitoringb
        Water                     Tissue
                                                                                    Special  Projects
U.S. FDA
(Seattle)
U.S. EPA
(Region 10)

Seattle-King  Co.
Health Dept.

Wash. Oept. of
Ecology

Metro
Total coliform bacteria (I)
Fecal conform bacteria (I)
Fecal streptococci (I,NR)
Total collfonn  bacteria (I)
Fecal conform  bacteria (I)
                             Fecal conform bacteria (I)(NR)
Total conform bacteria (I)
Fecal conform bacteria (I)

Fecal conform bacteria (I)   Fecal conform bacteria (I)
Vibrio spp.  (P)
C. perf
C. perfrlngens  (P)
Aeromonas  hyd
Yerslnia spp.
                                                                                    Aeromonas hydrophlla (I,P)
                                                                                               >.  (P)
                                                               Occasional  (most work is
                                                               1n fresh water)c
                                  C. perfrinqens  (I,P)
                                  E. coll (I)
                                  Vibrio parahaemolytlcus (P)
                                  Enterococci (I)
Wash.  Dept. of
Social  and
Health Services
Fecal  conform bacteria (I)
Fecal  conform bacteria (I)
Standard  plate counts (I)
  I - indicator of primary pathogens or specific pollution sources.  P  -  primary pathogen,
NR - not  routine.

b No agency  in Puget Sound conducts  routine monitoring of bacteria in sediments.

c Marine  studies have included  the fate and survival of  fecal colifonm bacteria and heterotrophic
bacteria  In  relation to dredging  operations.

-------
                                                       Microbiology Protocols
                                                               Future  Studies
                                                                 October 1986
                    RECOMMENDATIONS FOR FUTURE STUDIES
BACTERIAL INDICATORS

     Reliance on  a  single bacterial  indicator or  a  suite of indicators
may not be desirable as a general  rule  because the objectives  of different
studies may vary substantially.  For example, requirements for water pollution
monitoring differ  from those for evaluating  the acceptability  of shellfish
for human consumption.  In general, however, the following bacterial  indicators
should be considered for  use  in most  surveys  of bacterial  indicators in
Puget Sound:

     •    Fecal coliform bacteria

     t    Fecal coliform bacteria/E. coli

     •    Enterococci

     t    C.  perfringens.

These  indicators  might  be used  individually or  in various  combinations
depending upon the specific objectives of  each study.  The  best  uses and
the limitations of each indicator are discussed later in this chapter.

PRIMARY PATHOGENS

     In some cases, disease outbreaks may  require the direct investigation
and identification of primary pathogens  because  use of an  indicator is
inappropriate.   For example,  outbreaks of acute gastroenteritis may  show
no correspondence  between  the presence of  a  specific indicator, such as
fecal  coli form  bacteria and the primary pathogen,  such as Yersinia entero-
colitica (Hunger et al. 1980).  Direct  identification of primary pathogens
may also  be  appropriate during  reconnaissance surveys in  areas lacking
historical  data.   The choice of pathogen should be determined by the  nature
of the disease and by seasonal considerations.  For example,  Vibrio bacteria
generally are found in the  warmer months,  whereas Yersinia  bacteria are
more prevalent during colder periods of the  year.

-------
                                                       Microbiology Protocols
                                                      Sampling Considerations
                                                                 October  1986
                     SPECIAL SAMPLING CONSIDERATIONS
     Choice of an  optimum matrix for sampling  (e.g., water column, sediment,
interstitial water, or  tissue) will depend  upon the objectives of each
study.   For example, public health impacts  are most likely through ingestion
of contaminated shellfish or contact with  contaminated  water.  Shellfish
tissues or recreational waters may therefore be  the materials most appropriate
for analysis in studies focusing on  health risks.  The  identification of
present  point sources of  pollution can  best  be determined by analysis of
water samples, whereas  long-term pollution  trends may  best  be described
by sediment analysis.

WATER COLUMN

     Water  sampling  can result  in  highly variable  data because bacteria
are not uniformly  distributed throughout the  water column  (Gameson 1983)
and  sample volumes generally are limited to  50-100 mL.  One major cause
of spatial heterogeneity is the tendency for bacterial cells to concentrate
in a thin microlayer on the surface of  the  water.  Because bacterial abundances
in the microlayer may  exceed abundances  in  underlying  surface water by
several orders of  magnitude (Hardy 1982), it is  reconmended that the microlayer
and underlying water be sampled separately.   However, sampling of the microlayer
requires  specialized techniques that have  yet to be standardized.  Also,
collection and analysis of samples from both the microlayer  and underlying
water  at  each station may  be too expensive for many  routine monitoring
programs.  Thus, if  separate samples cannot be collected within the constraints
of a particular program, it is recommended that the microlayer be included
in the sample by  using  the traditional "scoop" method  of  surface water
sampling  (U.S. EPA 1978).   This method  involves  plunging an open  bottle
straight down to a depth  of 15-30 cm below  the water surface, moving it
horizontal  to the surface  while tipping it slightly  to  let  trapped air
escape, and removing the bottle in a vertical  position.   It  is recommended
that samples be collected using a wide-mouth (12-15 cm) bottle to facilitate
inclusion of the microlayer. •

SEDIMENTS

     Sediments  are  known to  be  heterogeneous with  respect to  types and
numbers of bacteria.  In addition, the bacteriological composition of sediments
may  have  little relationship to public health  impacts.  For these reasons,
routine monitoring of sediments in Puget Sound presently  is  not undertaken
by any organization (Table 1), and is  not recommended except under special
circumstances.

     One  special  application of sediment monitoring is analysis for C.  per-
fringens to trace the distribution of sewage.  Because spores of this bacterium

-------
                                                      Microbiology Protocols
                                                     Sampling Considerations
                                                               October 1986
are associated  with fecal pollution and  survive over long  periods, they
offer the advantage of providing a cumulative record of sewage  influence
suitable for  long-term monitoring surveys.   However, the  fact that the
spores are  persistent in the environment, and thus accumulate,  renders
analysis for C. perfringens  in sediments  inappropriate as a basis for providing
regulatory  guidelines.

TISSUE

     Sampling and analysis of shellfish tissue present fewer  problems than
sampling and analysis of water and  sediment (see APHA  1985a).   Shellfish
sampling is very important because  the  consumption of shellfish as food,
sometimes in the  raw state, may present a serious public  health  hazard.

     Shellfish offer  several  advantages  for sampling:   they concentrate
bacteria, can  be  sampled relatively  easily, and reflect pollution  levels
over  relatively long periods in both  sediment and water.  In Puget Sound
it is recommended that one  or several  shellfish species  of  recreational
or commercial  importance  be sampled routinely at each major  harvesting
area.  The  use of a  small number (preferably one) of species  as  standards
will  reduce  the variation among stations and sampling periods that  results
from interspecific differences in the propensity to concentrate  bacteria.
Because the whole organism is eaten,  the whole body should be prepared
for analysis.

-------
                                                        Microbiology  Protocols
                                                           Use  and  Limitations
                                                                 October 1986
          USES  AND  LIMITATIONS OF RECOMMENDED  BACTERIAL INDICATORS


     Recommended bacterial  indicators  for  monitoring  in  Puget  Sound are
 presented in Table 3 for each  kind of sample matrix.

 FECAL COLIFORM BACTERIA AND FECAL COLIFORM BACTERIA/E.  COLI

     The  density of fecal  col i form bacteria  has commonly  been used as an
 indicator of fecal pollution  and as an indicator of the presence of pathogens.
 The widespread use of fecal  col i form bacteria as an indicator has  the advantage
 of providing a basis for comparisons with historical  data.  However,  there
 are  several distinct, and often substantial,  limitations to using fecal
 coliform bacteria for these purposes.  In addtion, the  densities of colifomt
 bacteria  may  not  accurately  reflect public health risks  (Hanes and Fragala
 1967).  Recent research has indicated that, although many  species of non-patho-
 genic  enteric bacteria (e.g., coliform bacteria) are  viable in aquatic
 systems, they are not totally recoverable using  conventional  techniques
 (Roszak  et al.  1984;  Xu et al. 1982).   Normal culturing  techniques may
 seriously underestimate  the  concentrations of these organisms  in the environ-
 ment.   Because  the  degree of  recoverability may  also  depend on variable
 environmental factors (e.g., nutrient concentrations),  it  may be difficult
 to develop general  bacteriological standards that could apply to all areas.

     Cabelli  et  al. (1983) concluded that  fecal  coliform bacteria were
 inferior to enterococci as  indicators of the presence of pathogens  in marine
 recreational waters with respect to evaluating public health  risks.   Survival
 times for coliform bacteria  are  substantially less than  for many pathogens
 (Borrego  et al. 1983), complicating  efforts  to correlate counts of fecal
 coliform bacteria with  the densities  of  pathogens  at any specific  time.
 Another problem associated  with  the use of fecal  col i form bacteria as indicators
 is the fact that  they are not  specific to mammalian  fecal pollution.   For
 example,  the  fecal  coliform bacterium  Klebsiel la is common in pulp mill
 effluents.

     For all of  the above  reasons,  data on fecal coliform bacteria cannot
 in themselves be  considered  adequate  for  a thorough assessment of public
 health risks.  Fecal coliform bacteria continue to  be used as an indicator
 because other indicators also  have deficiencies, and because measurements
of fecal  coliform  bacteria provide a basis for comparisons with historical
data.  The lack  of  specificity of fecal  coliform  bacteria to mammalian
 fecal  pollution prompted  the development of  a membrane filtration method
 for enumerating  E. coli  (Dufour et al.  1981).  Unlike  fecal coliform bacteria
as a  group,  E.  coli  are specific to mammalian  fecal pollution.  The enumeration
method incorporates  a technique  for distinguishing  E. coli from other  fecal
coliform bacteria.   The result is notated  as fecal  coliform bacteria/ E. coli.
However, with respect specifically to recreational water quality criteria,


                                     8

-------
                 TABLE 3.  RECOMMENDED BACTERIAL  INDICATORS
                       FOR  MONITORING  IN PUGET SOUND
                                             Matrix
Objective
    Water*
   Sediment
    Tissue
Recreational use
evaluation (e.g.,
swimming)

Pollution monitoring
and/or water quality
surveys
Shellfish consump-
tion evaluation
Fecal conform
  bacteria
Enterococci

Fecal conform
  bacteria
L. coll
C^ perfringens
Enterococci

Fecal coliform
  bacteria
          D
Fecal  conform
  bacteria
C. perfringens
Fecal conform
  bacteria
E. coli
C. perfrinqens
                  Fecal conform
                   bacteria
a Fecal conform bacteria are recommended here because current water quality
standards are based on them.  If these standards  change to include only
enterococci, it may  still  be useful  to measure  fecal  coliform bacteria
to maintain continuity with historical databases.

b U.S. EPA and National Oceanic and Atmospheric Administration  (NOAA) currently
are co-sponsoring  research Into the  application of  the  enterococci and
E. coli indicators  for shellfish harvesting waters.

-------
                                                       Microbiology Protocols
                                                          Use  and Limitations
                                                                October 1986
U.S. EPA  has not recommended  the use of E. coli for marine  and estuarine
waters.

     Laboratory  analyses of fecal  coliform bacteria  are  conducted using
one of  two most probable number (MPN) techniques (i.e.,  EC  and  A-l) or
using a membrane filtration  (MF) technique.  The statistical  reliability
of the MF technique is greater than the MPN procedures (APHA 1985b).   However,
factors such as turbidity  may reduce counts for the MF technique (Berger
and Argaman 1983).  Thus, the results generated by the MF and MPN techniques
may not be directly comparable, and their inconsistent use by Puget Sound
organizations limits comparisons of data gathered by different  agencies.

ENTEROCOCCI

     Enterococci  are  streptococcus bacteria indigenous to the intestines
of warm-blooded animals.  U.S.  EPA recommends  their use as indicators of
fecal pollution in recreational waters because of the previously mentioned
limitations of fecal coliform bacteria analysis, and because of  the following
characteristics:

     •     Because  the concentration of enterococci has a greater corres-
          pondence to the incidence of gastroenteritis than do  concen-
          trations  of E. coli  or  fecal  coliform bacteria,  it is a
          better  indicator of public health risk in recreational  marine
          waters  (Cabelli et al. 1983).

     0     Enterococci may die  off more slowly in sediments  than fecal
          coliform bacteria (Van Donsel and Geldreich 1971), and therefore
          be better indicators  of sediment contamination.

     •     Because they are tolerant to  high salinity,  enterococci
          are of  particular value in analysis of marine  waters (Coler
          and Litsky 1976).

     0     Taxonomic identification of streptococcus bacteria  can  be
          undertaken easily (e.g., API biochemical strips) and, unlike
          the situation  with  fecal coliform bacteria,  can  reveal  the
          kinds of mammalian  pollution (e.g.,  humans, livestock).
          This advantage arises from the fact  that particular kinds
          of mammals harbor characteristic  species of  streptococcus
          bacteria (e.g., S.  bovis in cattle).

     0     Techniques of  gene  fingerprinting (DMA analyses) have been
          undertaken using  streptococcus bacteria and can more  positively
          link bacteria in the  environment to identical  bacteria found
          in source  effluents,  thereby confirming the source(s)  of
          contamination.  Although not suitable for routine  monitoring
          surveys, genetic analysis may be useful  in certain research
          applications.


                                    10

-------
                                                       Microbiology  Protocols
                                                          Use and  Limitations
                                                                October 1986
CLOSTRIDIUM PERFRINGENS

     C.  perfringens is consistently associated with fecal wastes and provides
a usable,  state-of-the-art  marker for delineating the deposition and/or
movement  of sewage participates that  is more reliable than the traditional
coliform bacteria  indicators (Emerson  and  Cabelli 1982).   C.  perfringens
is recommended for water,  sediment, and  tissue because it is present  in
wastewater at concentrations  of 10^-104  per 100 mL (Fujioka and Shizumura
1985), and because its resistance to chlorination and environmental factors
closely resembles  that of enteric viruses (Bisson and Cabelli 1980).
                                    11

-------
                                                         Mi,c.ro'bio,l qgy  Protocol s
                                                          ' La'bbratfory .Procedures
                                                                   October  1986
        LABORATORY  PROCEDURES  FOR  RECOMMENDED BACTER'lAL INDICATORS
     Recommended laboratory procedures for the bacterial indicators  listed
in Table 3 are summarized in Table 4.
                                      12

-------
              TABLE 4.  RECOMMENDED LABORATORY PROCEDURES FOR BACTERIAL INDICATORS
Test Organisms
                                                 Laboratory Procedures
      Water
             Sediment
      Tissue
Fecal collform
  bacteria
Fecal coliform
  bacteria/E. coll
Enterococci
C. perfringens*
MPN tubes using A-I
broth (APHA 1985b)
(fecal coliform
bacteria/100 mL)
        MPN tubes using  A-I
        broth (APHA 1985b)
        (fecal  coliform
        bacteria/100 mL)
mTEC (DuFour et
1981)
(E. coll/100 n»L)
al
mE (Levin et al.  1975)
(enterococci/100  mL)

MPN tubes using  iron
milk (St. John et al.
1982) (C. perfrin-
gens/100 mL)
        MPN tubes  using  iron
        milk (St.  John et  al.
        1982) (C.  perfrin-
        gens/100 g)
MPN tubes using EC
broth (APHA 1985a)
(fecal coliform
bacteria/100 mL)
MPN  tubes using  iroi
milk  (St.  John et al
1982) (C. perfrin-
gens/100 mL)
a Two  laboratory  techniques are available  for C. perfringens: mCP by membrane filtration foi
water  (Bisson and  Cabelli 1979) and  sediment (Emerson and Cabelli 1982), and. iron milk tube
using  MPN techniques  (St. John  et al.  1982).  The  latter method is recommended (pending an
comparative data) because the  procedure is simpler and less costly.
                                               13

-------
                                                        Microbiology Protocols
                                                                         QA/QC
                                                                  October 1986
                 QUALITY ASSURANCE/QUALITY CONTROL  (QA/QC)


     Laboratory and analytical QA/QC  procedures  are discussed in detail
in U.S. EPA (1978) and APHA (1985b).   Special problems  exist in microbiological
analyses because analytical  standards, known additions, and reference samples
generally are  not available.   However,  a minimum QA/QC program should include:

     •    Ten percent  of the total  number of samples analyzed in duplicate

     0    Ten percent  of the total  number of samples split and analyzed
          by  two or more laboratories

     •    Sterile distilled  water transported to the field, transferred
          to  a sample  bottle,  and processed routinely to ensure samples
          were not contaminated  during collection and transport

     •    Repeated sampling at one site  during  varying conditions
          (e.g., tides,  weather) to evaluate variability in the field.
                                    14

-------
                                                        Microbiology Protocols
                                                                    References
                                                                  October 1986
                                REFERENCES


APHA.  1985a.  Laboratory procedures for  the  examination  of seawater  and
shellfish.  15th ed.  American Public Health Association.  144 pp.

APHA.  1985b.  Standard methods for  the examination of water and wastewater.
16th ed.  American Public Health Association.  1268 pp.

Berger, P.S., and Y. Argaman.   1983.   Assessment of microbiology and turbidity
standards for drinking water.   EPA 570-9-83-001.  U.S.  Environmental Protection
Agency, Washington, DC.

Bisson,  J.W., and  V.J. Cabelli.   1979.   Membrane  filtration enumeration
method for Clostridium perfringens.  Appl. Environ. Microbiol. 37:55-66.

Bisson,  J.W., and  V.J. Cabelli.  1980.   Clostridium perfringens as a water
pollution indicator.  J. Water Pollut. Control Fed. 52:241.

Borrego,  J.J., F. Arrabal,  A.  de Vicente, L.F. Gomez, and P. Romero.   1983.
Study of microbial inactivation in the marine environment.  J. Water Pollut.
Control Fed. 55:297-302.

Cabelli,  V.J., A.P.  Dufour,  L.J.  McCabe, and M.A. Levin.   1983.  A marine
recreational water quality  criterion  consistent  with indicator concepts
and risk analysis.  J. Water Pollut. Control Fed. 55:1306-1314.

Coler, R.A.,  and  W. Litsky.   1976.  Pollutants  and aquatic ecosystems:
biological aspects of  water  quality  problems,  pp.  355-384.   In:  Industrial
Microbiology.   B.M. Miller and W.  Litsky  (eds).  McGraw-Hill  Book Co.,
New York, NY.

Dufour,  A.P., E.R. Strickland, and  V.J. Cabelli.  1981.   Membrane filter
method for enumerating Escherichia coli.   Appl. Environ.  Microbiol.  41:1152-
1158.

Emerson, D.J., and V.J. Cabelli.  1982.  Extraction of Clostridium perfringens
spores from bottom sediment samplers.  Appl. Environ. Microbiol. 44:1144-1149.

Fujioka, R.S. and  L.K. Shizumura.  1985.   Clostridium perfringens. a reliable
indicator of streamwa'ter quality.  J. Water Pollut. Control  Fed. 57:986-992.

Gameson,  A.L.N.   1983.  Investigations of sewage discharges to some British
coastal  waters. Water Resources Centre Technical  Report,  TR 193.  Bucks,
United Kingdom. 40 pp.
                                     15

-------
                                                        Microbiology Protocols
                                                                    References
                                                                  October  1986
Hardy,  J.T.  1982.   The  sea surface  microlayer:  biology,  chemistry,  and
anthropogenic enrichment.   Prog. Oceanogr. 11:307-328.

Hanes,  N.B., and R.  Fragala.  1967.   Effect of seawater concentration on
survival of indicator bacteria.  J. Water Pollut. Control  Fed.  39:97-104.

Levin, M.A., J.R.  Fischer,  and V.J. Cabelli.  1975.  Membrane filter technique
for enumeration of enterococci in marine waters.  Appl.  Microbiol.  30:66.

Munger,  F., T.F. Wetzler, A.A. Heyward, and R.J. Swartz.   1980.   Isolation
of Yersinia enterocolitica  from Saxidomus giganteus harvested  from Seattle
beaches.  Presented  at the  annual meeting of the American  Society  for  Micro-
biology.

Roszak, D.B., D.J. Grimes,  and R.R. Colwell.  1984.  Viable but nonrecoverable
stage of Salmonella  enteritidas in aquatic ecosystems.  Can. J. Microbiol.
30:334-338.

St. John,  E.W.,  J.R.  Matches, and  M.M. Wekell.   1982.   Use of iron milk
medium for enumeration of Clostridium perfringens.  Journ.  Assoc. Off.
Anal. Chem. 65(5):1129-1133.

U.S.  Environmental  Protection Agency.  1978.  Microbiological methods  for
monitoring the environment.  EPA-600/8-78-017.  U.S. Environmental  Protection
Agency, Cincinnati,  OH.  337  pp.

Van Donsel, D.J.,  and  E.E.  Geldreich.  1971.  Relationship  of salmonellae
to fecal coliforms in bottom  sediments.  Water Res. 5:1079-1087.

Xu, H., N.  Roberts,  F.L. Singleton, R.W. Atwell, D.J. Grimes, and R.R. Colwell.
1982.  Survival  and  viability of nonculturable Escherichia col i and Vibrio
cholerae in the estuarine and marine environment.  Microb.  Ecol. 8:313-323.
                                    16

-------