United States      Office of Research and    EPA/600/R-94/038d
         Environmental Protection   Development      April 1994
         Agency        Washington DC 20460
c/EPA    Quality Assurance
         Handbook for
         Air Pollution
         Measurement
         Systems

         Volume IV: Meteorological
         Measurements
         (As Revised August, 1989)

-------
                                       EPA-600/R-94/038d
    QUALITY ASSURANCE HANDBOOK

                     FOR

AIR POLLUTION MEASUREMENT SYSTEMS


    Volume IV — METEOROLOGICAL MEASUREMENTS

                as revised August. 1989
           U. S. ENVIRONMENTAL PROTECTION AGENCY

              Office of RcjMich sod Development
             Research Triangle Bark, North Carolina 27711


                                 JTTV.
                                   Printed on Recycled Paper

-------
                                                             Section No.    4.1
                                                             Revision No.    0
                                                             Date:    17 Sea 89
                                                             Page:     1  of 1
                                ACKNOWLEDGEMENTS
    This volume of the Quality Assurance Handbook has been prepared by Thooas
J. Lockhart, CCM.  of the Meteorological  Standards Institute on Fox Island.
Washington.  Sections 4.0.4.4, 4.1.8,  4.4.0,  4.5.0 and 4.6.0 were used fro*
EPA-60O/4-82-060 with only slight modification.  The contributions of the
authors of  this work, Peter L. Flnkelstein, Daniel A. Mazzarella, Thomas J.
Lockhart, William  J. King, and Joseph H. White are hereby gratefully
acknowledged.

     The first draft of this volume received  the contributions of the Critical
Review Panel who are:

                     Dr. C. Bruce Baker
                     Mr. Jack A. Bowen
                     Dr. Harold L. Crutcher. CCM
                     Prof. Gerald C. Gill, and
                     Mr. Alvin L. Morris.

    Work on this volume and the workshops which followed itspublication have
been Immeasurably  helped by the generous cooperation of the following companies
who made their products and services available  to the author:

     AeroVironment Inc.            Campbell Scientific, Inc.
     Cllmatronics  Corporation      Hollls International Limited
     Odessa Engineering            Rotronlc Instrument Corp.
     Qualimetrics, Inc.            Teledyne Geotech
     Waters Manufacturing. Inc.    R. M. Young  Company

    This revised  Volume  IV was produced under contract 68-02-4553. Mr. Jack  A.
3owen, Project Officer.
                                    DISCLAIMER

     Mention of trade names or commercial products does not constitute EPA
 endorsement or recommendation for  use.

-------
                                           Section No.  1
                                           Date February 24,  1994
                                           Page 1


            OVERVIEW OF THE  INTERIM EDITION OF VOLUME IV

     The Quality Assurance  (QA) Handbook is comprised of five
volumes:  Volume I (Principles), Volume II (Ambient Air Methods),
Volume III  (Stationary Source Methods), Volume IV (Meteorological
Measurements) ,  and Volume V (Precipitation Measurement Systems) .
Much of the material in Volumes II,  III and V are out-of-date and
some portions of these volumes have long been out-of-print.

     EPA is now preparing an updated version of the QA Handbook
series which will be available in September 1995.  To meet the
needs of the user community until the updated version is
available, EPA has published Interim Editions of Volumes I, II,
III, IV and V.   Each volume of the Interim Editions, is being
issued as a complete unit with out-of-date sections either
deleted or modified using addendum sheets and handwritten
notations in the text.

     This volume and the other four volumes of the Interim
2dition of the QA Handbook are available at no charge from:

               USEPA/ORD
               Center for Environmental Research Information
               26 West Martin Luther King Drive
               Cincinnati, Ohio  45268

     Since  this volume was updated in 1989, this interim edition
is being published without  changes.  The updated version, which
will be available in September 1995, will contain information on
-eteorological systems for measuring upper air meteorological
conditions.

     The user of the QA Handbook is cautioned to bear in mind
that the information provided  in the handbook is for guidance
purposes only.  EPA regulations are published in the Code of
Federal Regulations  (CFR),  When information in the CFR conflicts
with information in the QA  Handbook, the CFR shall be considered
the authoritative and legally  bonding document.
                              William J. Mitchell
                                   Chief
                         Quality Assurance Support Branch

-------
                                                             Section No.  4. ii
                                                             Revision No.    0
                                                              Date:   17 Sep 89
                                                              Page:    1  of  2
                                   VOLUME IV
                               TABLE OF CONTENTS
Section
4.1    ACKNOWLEDGEMENTS
4.11   TABLE OF CONTENTS

4.0    GENERAL PROGRAM REQUIREMENTS
       FOR METEOROLOGICAL MEASUREMENTS
     .0.0   OUTLINE, PURPOSE AND OVERVIEW
     .0.1   GLOSSARY
     .0.2   STATE OF THE ART
     .0.3   DATA REQUIREMENTS
     .0.4   MEASUREMENT REQUIREMENTS
     .0.5   REFERENCES
4.1
    4.
    4.
    4.

    4.
    4.
    4.
    4.
    4.
    4.
    4.
  0
 .1
1.2
GENERAL ASPECTS OF QUALITY ASSURANCE
FOR METEOROLOGICAL MEASUREMENTS
     OUTLINE AND SUMMARY
     PLANNING FOR A QUALITY ASSURANCE PROGRAM
     ORGANIZATION OF AUTHORITY AND
     RESPONSIBILITY
     QUALITY CONTROL GUIDELINES
     TRACEABILITY PROTOCOL
     ESTIMATING PRECISION AND ACCURACY
     SYSTEM AUDITS
     PERFORMANCE AUDITS
     DATA VALIDATION PROTOCOL
     QA REPORTS AND CORRECTIVE ACTION
     REFERENCES
  3
  4
  5
  6
  7
  8
  9
    4.1.10
4.2    QA FOR WIND SPEED.
       TURBULENCE
                    WIND DIRECTION AND
    4.2.0
    4.2.1
    4.2.2
    4.2.3
    4.2.4
    4.2.5
    4.2.6
    4.2.7
    4.2.8
    4.2.9
      OUTLINE AND SUMMARY
      TYPES OF INSTRUMENTS
      SPECIFICATIONS
      ACCEPTANCE TESTING
      INSTALLATION
      CALIBRATION
      OPERATIONS, MAINTENANCE AND QC
      PERFORMANCE AUDIT METHODS
      ESTIMATING ACCURACY AND PRECISION
      REFERENCES
4.3    QA FOR TEMPERATURE AND TEMPERATURE
       GRADIENTS (AT)
    4.3.0   OUTLINE AND SUMMARY
    4.3.1   TYPES OF INSTRUMENTS50
    4.3.2   SPECIFICATIONS
    4.3.3   ACCEPTANCE TESTING
    4.3.4   INSTALLATION
    4.3.5   CALIBRATION
    4.3.6   OPERATIONS, MAINTENANCE AND QC
                                                     Pages  Rev.  Date
 1
 2

25

 5
 2
 6
 3
 7
 2

33

 3
 3
 1

 2
 4
 6
 2
 1
 7
 2
 2

83

 4
 4
32
 5
 8
 4
 2
20
 1
 3
                                                       3
                                                       4
                                                       4
                                                       2
                                                       1
                                                       9
                                                       3
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89

-------
                                                             Section No.   4.11
                                                             Revision No.    0
                                                              Date:    17 Sep 89
                                                              Page:     2   of  2
Section
                                   VOLUME IV
                                TABLE OF CONTENTS
    4.3.7   PERFORMANCE AUDIT METHODS
    4.3.8   ESTIMATING ACCURACY AND PRECISION
    4.3.9   REFERENCES
4.4.0   QA FOR PRECIPITATION MEASUREMENTS

4.5.0   QA FOR RELATIVE HUMIDITY OR DEW POINT
        TEMPERATURE

4.6.0   QA FOR SOLAR RADIATION MEASUREMENTS

4.7.0   QA FOR ATMOSPHERIC PRESSURE
Pages  Rev.   Date
8
1
1
7
10
8
2
0
0
0
0
0
0
0
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89
9/17/89

-------
                                                             Section No.  4.0.0
                                                             Revision No.     0
                                                             Date:    17 Sep 89
                                                             Page:     1  of  5


                                  Section 4.0
                       GENERAL PROGRAM REQUIREMENTS FOR
                          METEOROLOGICAL MEASUREMENTS
                                    OUTLINE
Section                                                      Pages Rev.  Date

4.0.0  OUTLINE.  PURPOSE AND OVERVIEW
       OF THE QUALITY ASSURANCE HANDBOOK                     5     0     9/89

4.0.1  GLOSSARY                                              2     0     9/89

4.0.2  STATE OF THE ART                                      60     9/89

       4.0.2.1  AUDITOR SURVEY
       4.0.2.2  INTERVIEW SUMMARY

4.0.3  DATA REQUIREMENTS                                     3     0     9/89

       4.0.3.1  REGULATORY PROGRAMS
       4.0.3.2  RESEARCH PROGRAMS
       4.0.3.3  CONTINGENCY PROGRAMS

4.0.4  MEASUREMENT REQUIREMENTS                              7     0     9/89

       4.0.4.1  MEASUREMENT SYSTEM
       4.0.4.2  DOCUMENTATION
       4.0.4.3  MOUNTING
       4.0.4.4  SITING

4.0.5  REFERENCES                                            2     0     9/89

-------
                                                             Section No.  4.0.0
                                                             Revision No.     0
                                                             Date:   17 Sep 89
                                                             Page:    2  of  5


                      PURPOSE AND OVERVIEW OF THE QUALITY
                               ASSURANCE HANDBOOK

    The purpose of this  volume  of the  QA Handbook is to provide information and
guidance for both the meteorologist  and the non-meteorologist who must make
judgments about the validity of data and accuracy of measurement systems.   Care
has been taken to provide  definitions  to help those making these Judgments to
communicate without ambiguity.  Methods are described in the handbook which
will objectively define  the quality  of measurements so the non- meteorologist
can communicate with the meteorologist or environmental scientist or engineer
with precision of meaning.

    The first section of the handbook  contains a special glossary of terms
necessary to meteorology and quality assurance.  Following that is an analysis
of the state of the art  from information and interviews of those practicing QA
in the air quality field.   The  final parts of the first section define some of
the requirements for gathering  data  which a QA effort can compare to the
practice of acquiring data.

    The second section  is  devoted to quality assurance and quality control as
it is applied to meteorological problems.  This section is somewhat independent
of the variable being measured.  Where the variable is important it is treated
individually.

    The final six sections are  variable-specific.  The most important wind
measurement is covered  in  considerable detail.  The temperature measurement
section concentrates on the temperature difference measurement used for
stability determination.  The final  four sections cover to an adequate depth
the measurement of humidity, radiation, precipitation and surface air pressure.
Examples are given where possible to help explain the methods and problems to
be found in programs of collecting meteorological data and assessing data
validity.

    The need for common understanding  is critical for the practice of quality
control  (QC) and quality assurance  (QA). This is achieved in part by the
definitions of the language used  within the discipline.  From that vocabulary,
the details of the systems and  procedures are defined in terms of the necessary
goals.

    There are a variety of QA/QC  definitions in  the literature and in common
usage.  Volume I. Section  No. 1.3 and  Appendix A provide some general
definitions.  Section  1.4  shows how  the elements of QA are distributed and
where in the section they  are described. The well known "quality assurance
wheel" is shown in Figure  1.4.1.  The  following discussion of definitions is
broader based to include meteorological requirements and explicit between QA
acd QC.

    The structure shown in Figure 4.0.0.1 below  is from ANSI/ASQC Q90-1987;
American National Standard, Quality  Management and Quality Assurance Standards
Guidelines for Selection and Use. The definitions in the glossary (4.0.1) and
the following descriptions are  structured  to fit Figure 4.0.0.1 and the
practices of meteorological measurement.

-------
                                                            Section No. 4.0.0
                                                            Revision No.    0
                                                            Date:   17 Sep 89
                                                            Page:    3  of  5
   (Organizational)
   V Structure  /
    / Confidence \
    I  To The   )
    ^Management/
      QUALITY MANAGEMENT ASPECTS'

     —^QUALITY. SYSTEM\
       s     * ~       * fc     ^.
                 QUALITY
                 CONTROL
                 ASPECTS-
          INTERNAL
           QUALITY
      \  ASSURANCE
»    *   ***+***• " .•»**««     _^%i ••  ' » O SX«^»^t«'^.NV
\  •.ASPECT>^|i»

  v    '•    ^.-:r^PlP
            :• 1 A'-~  -*• ,w 's.vXv\*v-V _;<$«$WvWX-;:
                                                           (Policy)
                                                         / Operational \
                                                         I Techniques I
                                                             Activities/
                                           (When Required
                                            By A Contract)
                                         EXTERNAL QUALITY
                                        ASSURANCE ASPECTS
                                                        L
                                                         ]
                                                          CONFIDENCE TO
                                                            PURCHASER
Figure 4.0.0.1
The Structure Of Quality (reprinted with permission
from the  American Society  for Quality Control)
    This figure will be described as it applies to a meteorological company
which manufacturers instruments and provides a variety of services.  It can
also apply to government organizations and temporary project-oriented
cooperative organizations.  At the top of the figure and the top of any
organized effort toward quality are the QUALITY MANAGEMENT ASPECTS.  These
aspects comprise the policy statement for the organization expressed by
management.  The statement is in writing as a company policy and signed by the
president of the company so that there will be no misunderstanding or confusion
about the quality goals of instruments so that they do produce valid data.   A
performance audit, then, is a challenge both to the instrument and to the
operator to independently verify that the measurement system is "in control."
Just as with system audits, the auditor is primarily a teacher and trainer.
The audit method should be the best possible method.  The operator should be
encouraged to upgrade the calibration methods to do a better job.

    Often the operator has no experience with meteorological instruments.
Often they are well qualified instrument technicians, but the auditor  is an
expert, or should be.  A mutually valuable goal is for the operator  to  learn
what methods are necessary and most complete and adopt those for  the
calibration procedure.  When the instruments are all working perfectly  it  is
because they are getting the experienced attention it  takes for  "in  control"
operation.  At this point, the audit becomes a spot checking operation
producing documentation from an independent  individual verifying  this  "in
control" operation.

    If some basic ground rules are  followed,  the audit  is maximized  as  a
learning exercise.  One rule is that  the operator does all  the  instrument
handling.  There  is a general reluctance  to  handle unfamiliar  instruments.
They might get broken or changed in some mysterious way.  The  way to become

-------
                                                              Section No.  4.0.0
                                                              Revision No.     0
                                                              Date:    17 Sep 89
                                                              Page:     4  of  5


familiar  with them is to work with them.  The safest environment in which to
gain this experience is in the presence of someone who is familiar with the
instruments.

    Good  audit methods are as close to laboratory tests as a field site and the
ingenuity of the auditor will allow.  This inevitably requires a tower mounted
instrument to be taken down, complete with cable or substitute cable, so that
tests can be made in a physically convenient environment.  For wind sensors,
the bearing condition is of vital importance.  This can be measured only when
the sensor can be connected to the torque instrument with complete control and
freedom to move.  It is not a proper test to try on a tower or mast.
Performance audits, in exactly the same way as calibrations, challenge parts of
the system.  Ideally, one wants  to challenge all of the system, but that is
often  impossible.  Known free atmospheres are not available from NBS.
Controlled atmospheres like a wind tunnel or a thermal chamber or a "sun" lamp
only challenge a part of the system.  They leave out or drastically change the
important coupling function.  Even controlled atmosphere devices, such as a
wind tunnel, are not available to the auditor in the field.  All one can do is
impose  a known condition such as the rate of rotation for an anemometer, and
measure the system response.  This type of audit assumes that the
manufacturer's generic transfer  function applies to this sensor, or assume that
earlier wind tunnel tests of this sensor still apply (a good assumption if the
sensor  is not damaged).

    Another method is the ASTM collocated transfer standard method.  This is
 the most complete method from the standpoint of total system error sources but
 it has  two drawbacks.  First, it is limited to the conditions that prevail
during  the audit.  Secondly, it  is very sensitive to exposure or siting bias.
 It requires careful guidelines pointing out potential bias sources and ways to
watch  for them in the data.  These are covered in the variable-specific
 sections.

    A  performance audit program  using experienced independent auditors, whether
 internal or external to the organization, is the first step toward establishing
_a quality plan if one does not already exist.  The goal of the measurement
program is to have documented data.  The performance audit will point out areas
 required to get  the system "in control."  The auditor can help implement the
establishment of a quality system, or its key elements, in order to achieve the
necessary on-going activities to keep the measurement system "in control"
 continuously.

    The survey which led off the work of revising this handbook exposed some
 confusion  in the community of meteorological auditors with regard to the
 difference between performance audits and calibrations.  A form letter was
 ccoposed to discuss these differences and to ask for new numbers of audits
 conducted.  The  letter used  the  definitions found in the glossary  (4.0.1) and
 expanded on them with examples.  The principal difference  is independence of
 responsibility.  Some organizations perceived the documentation of the
 cccdition of the system "as  found" as a performance audit and the adjustment of
 the  system to acceptable operating conditions, documented  "as left," as a
 calibration.  Thus, a single  individual could both audit and calibrate during
 the  same visit.  By any accepted standard of quality systems definition, this
 whole process of testing and adjustment  is a calibration.  This properly
 documented calibration is the basis for claims of data validity.  All the

-------
                                                             Section No.  4.0.0
                                                             Revision No.     0
                                                             Date:    17 Sep 89
                                                             Page:     5  of  5

performance audit adds is an independent assurance that the calibrations were
thoroughly done and that the documents are complete and accurate.   Such
assurance must be entirely free of potential influence.

    The letter described situations where a single company can be structured to
provide both calibration and auditing services,  but cautioned that the
independence of the auditor requires a management structure insulating the
auditor from the budgetary concerns of the operating organization.   Responses
to this letter were few and in no case was the distinction challenged.    All
agreed with the concept of independence for QA audits.  Regulators should
acknowledge the distinction and require true independence.

    Situations arise where the fundamental principle of independence between
calibration and audit services is difficult to follow.  Small agencies may not
be able to contract for independent audits.  In the interest of documented
validity of data for all parties, innovative arrangements among different
agencies should be promoted.  The individual who operates and calibrates
instruments for Agency A might be asked to audit the Agency B instruments in
exchange for the operator at Agency B auditing the Agency A instruments.   This
practice would have the further benefit of stimulating communication about and
standardization of good audit methods.

-------
                                                             Section No. 4.0.1
                                                             Revision No.    0
                                                             Date:   17X Sep 89
                                                             Page:    1  of  2
4.0.1       GLOSSARY FOR METEOROLOGY AND QA/QC
ACCURACY - Is the degree of agreement of a measurement (or an average of
measurements of the same thing). X.with an accepted reference or true value, T,
usually expressed as the difference between the two values, X - T, or the
difference as a percentage of the reference or true value, 100(X-T)/T. or
sometimes expressed as a ratio, X/T.

CALIBRATION - Is a MEASURE of conformance to or discrepancy from a
specification or set of criteria for an instrument or system if necessary and
an ADJUSTMENT of the instrument or system to conform to the specification or
criteria.  A calibration may be performed by a person or agency within the
operating organization.

DAMPING RATIO (T)) - The damping ratio is calculated from the overshoot ratio
(Q). (21
                              In
DFLAY DISTANCE  (D) - The distance the air flows past a wind vane during the
    t it takes the vane to  return to 50 percent of  the Initial displacement.  (21

EXTERNAL QUALITY ASSURANCE -  is the activity designed to provide the purchaser
with confidence in the quality of what Is being purchased.

INTERNAL QUALITY ASSURANCE -  is the activity designed to provide management
with confidence that the quality system  is operating and the management policy
is  being carried out.                                    ">

INVERSION  (+AT) - is the inverted lapse  rate or an increase of  air
teaperature with height.   There is no general  limit for Inversion  strength.

LAPSE RATE (-AT) - is the  normal decrease of air  temperature with  height
Halted by the  auto convection rate of   3.4°C/100  m.
OVERSHOOT  (Q) - The ratio  of  the amplitude of  two  successive deflections of a
wind vane  as  it oscillates about the equilibrium position  after release from an
offset position of ten degrees, as expressed by the equation
                           e
                            n

 where 8  and 9     are  the amplitudes of  the n and n+1 deflections,
       n
 respectively.

-------
                                                             Section No. 4.0. 1
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    2  of  2


PERFORMANCE AUDIT - is a  report of conformance  to  or discrepancy from a
specification or set of criteria determined by   a  person or agency separate
from and independent of the  operating  organization.

PRECISION -  is the standard deviation of  a series of  measured values, X^,

about the mean measured value, X.  (see 4.1.5.1)

QUALITY ASSURANCE - All those planned  and  systematic actions necessary  to
provide adequate confidence  that a product or service  will satisfy given
requirements for quality,  [il

QUALITY CONTROL - The operational  techniques  and activities that are used to
fulfill requirements for  quality,  til

QUALITY MANAGEMENT - That aspect of  the overall management function that
determines and implements the quality  policy, [i]

QUALITY POLICY - The overall quality intentions and direction of an
organization as regards quality, as  formally  expressed by  top management. (11

QUALITY SYSTEM - The organizational  structure,  responsibilities, procedures,
processes, and resources  for implementing quality management,  til

REPRESENTATIVENESS -  is  the extent to  which a set of  measurements  taken in  a
space-time domain reflects the actual  conditions in the same or  different
space-time domain taken  on a scale appropriate  for a  specific application.  [4]

STARTING THRESHOLD  (S  ,  m/s) - The lowest speed at which a vane  will  turn to
        o             °                                                  o
within 5  of 8   (the  true direction) from an  initial  displacement  of  10 . [21


STARTING THRESHOLD  (U  ,  m/s) - The lowest speed at which a rotating anemometer
                      o
starts and continues  to  turn and produce a measurable signal when mounted in
its  normal position.  [31

HI  ANSI/ASQC,  1987a:   Quality Management and Quality Assurance
     Standards  -  Guidelines for Selection and Use.  ANSI/ASQC Q90-1987.  American
     Society  for  Quality Control.  Milwaukee,  WI  53203.

 [21  ASTM,  1985b:  Standard Test Method for DETERMINING THE DYNAMIC PERFORMANCE
     OF A WIND  VANE.  (Draft 8 of D22.ll) Amer. Soc. for  Testing  and Materials,
     Philadelphia.  PA 19103.

 [3]  ASTM,  1985a:   Standard Test Method for DETERMINING THE PERFORMANCE OF A CUP
     ANEMOMETER OR PROPELLER ANEMOMETER. (Draft 6 of D22.ll)  Amer.  Soc.  for
     Testing  and Materials, Philadelphia, PA  19103.

-------
                                                             Section No. 4.0.2
                                                             Revision    No. 0
                                                             Date:   17 Sep 89
                                                             Page:    1  of  6
4.0.2  STATE OF THE ART
    The achievement of predicted quality for a product or service can be the
delegated responsibility of an identifiable part of an organization.  The
practice of elevating quality to a management staff level is relatively new.
The value of quality control and the umbrella management structure of quality
assurance became clear when products, purchased against a specification, were
rejected by the purchaser.  When the cost of rework or scrap absorbs the
profit, an alternative will be found.  The alternative is to do it right the
first time and the path to that goal Involves training, in-process Inspection,
fLnal Inspection and all of the other QA functions designed to minimize scrap
and rework.

    The QA profession grew during World War II and thereafter as the U.S.
Government became a significant purchaser using comprehensivespecificatlons,
like the well known Mil-Specs.  In the '60sand '70s the practice of planned
obsolescence and using the customer as the final Inspector set up our
industries for failure against foreign competition with higher quality
standards.  The successful foreign producers, using the quality principles
developed in the United States, caused a resurgence of quality awareness.

    The Environmental Protection Agency (EPA) recognized the need to set
standards, develop standard methods and materials, and produce a system of
quality assurance to support validity claims for the data being collected in
response to the Clean Air Act.  In 1976 a Quality Assurance Handbook for Air
Pollution Measurement Systems: Volume I. Principles was published (EPA. 1976).
In 1977. Volume II. Ambient Air Specific Methods (EPA, 1977a) and Volume III.
Stationary Source Specific Methods (EPA, 1977b) were published.  This program
addressed the Criteria Pollutants which were covered by federal law.
Meteorological measurements were recognized as supportive to the Criteria
Pollutant measurement program but they were secondary.

    When the Clean Air Act was amended, permission for growth of source
strength (and thereby growth of industry) was granted as a con- sequence of
diffusion model predictions based on input meteorological data.  Now, the law
recognized the requirement for valid and represent- atlve meteorological data
and the need for a structure to provide documented assurance of validity.   In
1983, Volume IV. Meteorological Measurements  (EPA, 1983) was added  to the
Handbook family.

    Almost all of the QA work provided by the private sector was geared to air
and source chemistry.  When meteorology was added to the technical  requirements
list, a variety of solutions were applied by a variety of Individuals with a
variety of technical backgrounds.  The original Volume IV. was  like a
background guidebook for  taking meteorological measurements and general
suggestions for how QA and QC might be applied to the requirement for valid
da-.a.  This revision of Volume  IV. is  intended to be more specific  and more
informative and more in the spirit of  the other three volumes.  It  has not, and
cannot as yet, specify standard methods.  A greater success with predictive
models is necessary before knowledge will exist which can dictate the standard
me-hods to assure valid input data.  The premise of this Volume IV. is  that
measurements worth  taking are worth  taking right.

-------
                                                             Section No.  4.0.2
                                                             Revision    No.  0
                                                             Date:   17 Sep 89
                                                             Page:    2  of  6
       4.0.2.1 Auditor Survey

       This handbook is intended to document the methods currently in use in
Meteorological QA/QC and to point to methods which are optimum for meeting the
requirements suggested or defined in various EPA publications.  A starting
point toward this goal is a survey of all those active in performance and
system auditing of meteorological measurement programs.  Figure 4.0.2.1 is a
copy of the survey form sent to as many people with experience in auditing as
could be found.  The initial list, shown by company and location In Table
4.0.2.1, grew considerably with help from all the EPA Regions and many state
and local agencies.  The number of survey forms returned from each company is
also shown.

                    Table 4.0.2.1 -  Original Survey List
                  COMPANY
      * AeroVironment,  Inc.
        Dames & Moore
        Desert Research Institute
      * Enviro. Monitoring & Services  Inc.
        Environmental Research & Tech.
      * Environmental Research & Tech.
        Galson Technical Services,  Inc.
        Meteorological  Standards Inst.
        Research and Evaluation Assoc.
      * Research Triangle Institute
        Roy F. Weston,  Inc.
        RTF Associates
        Technical Environmental Enter.
        Tennessee Valley Authority
      * TRC Environmental Consultants
  CITY/STATE      NUMBER

Monrovia, CA         4
Atlanta, GA          2
Reno. NV             2
Thousand Oaks, CA    1
Fort Collins, CO     1
Concord, MA          4
E. Syracuse, NY      1
Fox Island, WA       1
Chapel Hill. NC
RTP, NC              1
West Chester, PA     1
Denver, CO           1
Aurora, CO           1
Muscle Shoals, AL    1
E. Hartford, CT
         • indicates companies  chosen for  in-depth  interview
        Of the 70 or so  forms  sent originally or copied and distributed within
an organization. 49 forms were returned.  The summary of these responses is
shown numerically on Figure  4.0.2.1.  The number of audits represented by the
survey is 12.195, where  the  definition of an audit is the challenge of one
instrument measuring a meteorological variable.  Each respondent was asked to
qualify himself by specialty,  using  three or more if necessary but indicating a
priority of  1,2 or 3.  Some  managers reported for their organization of
auditors.  The responses to  the questions were not weighted by numbers of
audits.  As  with most surveys,  a few points are useful but action should not be
based on the survey results.   Of the 49  survey forms returned with data, 21
came from the original  list. 9 came  from local, state or federal agencies and
19 came from others.  Of this  19, 5  came from utilities in the Northeast U.S.
(3 from Pennsylvania Power & Light)  showing a close relationship to Regional
Meteorologists and interest  in QA/QC in  the area.

-------
                                                       Section No.  4.0.2
                                                       Revision    No. 0
                                                       Date:   17 Sep 89
                                                       Page:    3  of  6
                 Survey of Meteorological  Measurement QC/QA  People
                        (please print or type your responses)
      NAME	COMPANY	
      ADDRESS	ADDRESS	
      PHONE  <     >                             «      >     .   _  .     „-,__
      I am a   meteorologist,    chemist*  	environmental  scientist,  	QA/QC
      professional,  	instrument technician*  	electronics tech..,  	engineer*
        •net.tech.,  ..modeler.  	manager,  _da€a analyst,    field hand,  (use
      T72.3 if you  are more than one)  This information wTTl  be summarized
      without the use of names or companies or agencies so please be candid.
         Consider an audit or  challenge as a QA observation of  the response
      of an instrument to a known input and consider  a calibration as an
      OPERATIONAL testing and  adjustment* as necessary, of an instrument.
      1. If one meteorological audit is defined as a  challenge  to one
         variable or one variable of a system, about  how many audits have you
         performed  in 198O-1984 	, 1985 	, 1986 	, 1987	?
      2. Did you34 usually jL&omctime* ^.never use a written procedure?
      3. If a calibration is defined as the testing and adjustment of one
         variable or one variable in a system, how many calibrations have you
         performed  in 198O-1984 	, 1985	* 1986 	, 1987 	?
      4. Did veu£9'usually / JTsometimes ./."ever use the manufacturer's
         calibration procedure?
      5. When you perform an audit, do you require the operator to remove the
         sensors from their mounted position?  31 Yes  *3£No  /^.Sometimes
      6. Do you require the operator to re-connect the sensor to the system
         when it is presented for audit? 2£Lves  {j^No
      7. If 6. is yes, is the re-connection made with££the operational
         cable, _Q.a substitute cable or  y either?
      8. Do you jP(7usual 1 v JJ_sometimes J!£never measure the starting torque
         of each anemometer bearing assembly and transducer?
      9. Do you l3L"sually /Qsometimes /£never n»«*ure *"« starting torque
         of each wind vane bearing assembly and transducer?
      1O. Do you ^.usually f^sometimesgfnever use the collocated transfer
         standard method for auditing a wind instrument?
      11. Do you3JL"sually ./^sometimes .^.never find the audited  instrument
         meets the  required specification?
      12. Do you challenge anemometers with known rates of rotation?
         /j>No - If yes, how many speeds 	?   	Synchronous or 	measured?
      13. Do you challenge direction vanes with a dividing wheel? g^Yes
         2J_No - If yes, how many angles 	?   	CM,	CCW, 	both.
      14. Will you fill out a more detailed questionaire  as  a contribution to
         the quality of this project? flSOYes
Figure 4.0.2.i  Survey Form and  Response Summary

-------
                                                             Section No. 4.0.2
                                                             Revision    No. 0
                                                             Date:   17 Sep 89
                                                             Page:    4  of  6
        The number of  audits,  sorted by the  technical background rated by the
respondent as  #1. is shown in Table 4.0.2.2.   It  is  comforting  to note that the
largest number of auditors consider themselves meteorologists first.  While the
largest number of audits were reported by persons considering themselves
managers first,  it is  likely that those they managed were distributed like the
rest of the group.  One organization reported by  a manager  listed 3.600 audits.
It is likely that the  discipline of the person doing most of the performance
audits is meteorologist.

       The information from each question gives some feeling about how the
audits reported  were conducted.   Question 2  shows that 76%  of the auditors back
their work with  a written procedure.  Question 4  shows that 64% of the auditors
usually use the  manufacturer's calibration procedure.  This answer pertains to
the calibration  function which most auditors perform as  a separate part of
their Job.

                         Table 4.0.2.2  - Survey Summary
Meteorological Performance Audits
#1 Specialty Number
Meteorologist 11
Engineer 8
QA/QC 7
Instrument Tech. 5
Manager 4
Environ. Sci. 4
Electronics Tech. 4
Chemist 3
Data Analyst 1
No indication 2
Total 49
Average number per year
Percent change
1980-84
1,291
253
387
129
2,115
224
510
352
0
0
5.261
1.052
	
1985
647
194
193
702
543
70
165
220
0
0
2,102
2,102
100
1986
473
188
237
102
551
130
181
256
0
30
2,256
2,256
7
1987
623
206
318
60
644
91
212
192
0
30
2,576
3,435
52
Total
3.034
841
1,135
669
3.853
515
1.068
1,020
0
60
12,195


        In question 5,  43% of the auditors either do not physically inspect  the
 sensor  or do so by performing the operator's function of climbing the  tower and
 renoving  the sensor.   Volume IV. should reduce that percentage to zero.
 Question  6 suggests that most auditors (67%) do both a physical and an
 operational challenge of the sensor when it is down from the tower.  The
 conditional question 7 shows a preference for the operational cable (71%) over
 a substitute cable.

        Questions 8 and 9 show only 43% of the auditors usually measure the
 s-arting  torque of the anemometer and only 41% usually measure the direction
 vane starting torque.   It looks like when an auditor decides to make this
 measurement,  both sensors are included.  Several respondents answered  "never"
 but indicated that they were getting equipment to make the measurement in the
 future.   Other' auditors inspect the bearing assemblies with educated fingers
 which tell the auditor whether or not they are "all right" but fail to provide
 numerical or objective documentation.

-------
                                                             Section No.  4.0.2
                                                             Revision    No. 0
                                                             Date:    17 Sep 89
                                                             Page:     5  of  6
       Question 10 shows that only 19% rely on the collocated transfer standard
(CIS)  method for auditing wind instruments.  Volume IV.  should help Increase
that number.   Question 11 may mean that 70% of the instruments audited are
working within specification, or it may mean that the audit methods used are
not rigorous enough to find the discrepancies.  The fact that half the audits
do not include a torque measurement, the only method short of a wind tunnel to
challenge starting threshold, points to the latter possibility.  Volume IV.
should help to Improve audit methods toward a standard practice so that this
question, asked in the future, will provide an unambiguous answer.

       Questions 12 and 13 show a difference in challenging speed and
direction.  There -were 73% of the auditors who indicated the use of a simulated
speed  to challenge an anemometer.  The most common number of speeds was two
(52X). followed by three (21%). then one (15%). and finally four or more (12%).
There  were 86% who Indicated a synchronous motor was used.  This near unanimity
is probably because of the availability of synchronous motors and the lack of
availability of simple measurement systems.  The measured method is the only
choice where good commercial power is not available.

       The direction challenges were not as uniform.  Of the 23 who indicated
the number of angles used, seven said 4, five said 6, four said 8, three said
5. and one each said 1. 12,  16 and 18.  There is no consensus there.  All but
two said they used both clockwise and counterclockwise rotation.  The two used
clockwise.

       If the survey did one thing, it demonstrated the need for guidance
toward an acceptable standard of performance auditing.  It also demonstrated a
recognition of need to move  toward that goal and a willingness to help in  the
process.  Only one of 47 said no to question  14.

       4.0.2.2 Interview Summary

       After the survey results were in, a series of visits was planned to talk
to private sector organizations which had a recognized role in quality
assurance of meteorological  measurements.  The first organization visited was
AeroVlronment. Inc. of Monrovia, California.  The half-day discussion with four
AV auditors was a frank exchange of methods currently in use, shortcomings of
Volume  IV and suggestions for  the content of  the revised Volume IV.  The
principle of starting torque measurements of  anemometer and wind vane shafts as
a field substitute for starting  threshold wind speed determination in a wind
tunoel was accepted.  The principle of operators doing all the climbing or
handling of sensors was currently practiced.

       The second interview  was  at Environmental Monitoring and Systems,  Inc.
in Thousand Oaks, California.  Half-day discussions with  two meteorological
auditors reinforced the belief that some organizations were advanced in the
practice of meteorological QA.   Comprehensively written audit procedures were
followed.  Questions of the  difference between an audit and a calibration  were
correctly answered with authority.  The need  for uniform expectations or
requirements was expressed  in  the context  of  competitive  bidding  for providing
audit services.   It was felt that the new  Volume IV could help buyers of

-------
                                                             Section No.  4.0.2
                                                             Revision    No. 0
                                                             Date:   17 Sep 89
                                                             Page:    6  of


services to specify a scope of work  in enough detail to both assure a
comprehensive service and provide a  fair bidding competition.

       The third interview was with  Environmental Research and Technology (now
ENSR) in Concord, Massachusetts, perhaps the largest of the five organizations
in terms of meteorological services  and auditing.  Some different concepts and
practices were found, particularly in the area of starting threshold
determination.  Seven meteorologists, field auditors and QA specialists were
present during the half-day discussion.  This organization was a leader in the
field of providing meteorological monitoring services to industry.  As a result
of the history of providing all  services including design, installation.
operation, data summarization and QA auditing, an interesting discussion was
held on the subject of  independence  between operators/calibrators and auditors.

       The fourth interview was  with the head of th> field operations
department of TRC Environmental  Consultants in "E:'Hartford, Connecticut.  This
organization was also a leader  in providing full meteorological monitoring
services.  Their procedures developed in a different way.  They calibrated
their sensors by wind tunnel  testing in their calibration facility and employed
a regular replacement of sensors in  the field.  All of the performance auditing
related to sensors was  done by QA personnel in the calibration facility.  This
nethod requires a spare set of  sensors be available for each client.  The
nethods described in Volume  IV  for calibrating or auditing in the field are not
necessary if you have a wind  tunnel  and employ the interchangeable sensor
nethod.

       The final interview was  with  a meteorologist/auditor from  the Research
Triangle  Institute of Research  Triangle Park, North Carolina.  Since two of the
original Volume  IV authors'were at RTI when the work was written, it was not
surprising to find the  methods  employed  to be acceptable standard methods.  The
level of quality of  the field standards used  in auditing was the  highest, as  it
vas with most of these  organizations.

       This  series of  interviews provided valuable  insights and confirmations
about the best methods  to use for  meteorological quality assurance practices.
It showed the field  to  be well  practiced  at the  level of the largest and best
consulting organizations.   The  task  for Volume  IV  is  to provide a basis for a
standard practice  in this field at  all levels, and  to provide a measure by
which those  practicing  in the field  can be judged by  those with the final
authority to accept  or  reject data on the basis of documented validity.

-------
                                                             Section No.  4.0.3
                                                             Revision No.     0
                                                             Date:    17 Sep 89
                                                             Page:     1  of  3
4.0.3  DATA REQUIREMENTS
    There are a variety of reasons why meteorological data are collected.  Some
reasons relate to regulatory requirements or national monitoring programs.
Some data are collected for the purpose of research.   Some data are collected
against the contingency that they may be needed at some future time.  Sometimes
data are collected for one reason and then used for other reasons.

    The philosophy upon which this volume rests is the belief that data need to
have an estimation of uncertainty before the numbers can be dignified by the
title "data."  The estimation might be a simple declaration such as "The
meteorological measurement program was operated in conformance with PSD
guidelines."  This cites the accuracy requirements for PSD as the uncertainty
level for the data and promises that the documentation required for validity
claims for a PSD application will be available to back these data.  When such
an estimation exists and rests on documentation of performance, decisions can
be made as to whether or not these data are appropriate for the application.

    4.0.3.1 Regulatory Programs

            4.0.3.1.1  PSD
            The regulatory program used in this document, and to some extent in
the On-Site Meteorological Program Guidance for Regulatory Modeling
Applications (EPA, 1987b). is the Ambient Monitoring Guidelines for Prevention
of Significant Deterioration (PSD) (EPA, 1987a).  This is the most explicit
guideline and one requiring a quality of instrument performance available only
from sensitive instruments.  Recently it seems to be used for other programs as
a "standard" of specification.

            Most air quality dispersion models described in the Guideline on Air
Quality Models (EPA, 1986) used for PSD applications are Gaussian models
requiring input data which represent the conditions at the site of interest and
which follow a prescribed data content and format.  The models require five
meteorological inputs.  They are:

            1) Wind speed - representing the average wind speed at 10 m above
the ground (and additional heights for elevated sources) during each hour
calculated by a scalar average or mean of samples taken during the hour,
usually in 15 minute increments.  The samples may be the integrated wind run
during the sample period  (one or two seconds is often used) or instantaneous
samples of speed.  A resultant vector magnitude does not represent the initial
dilution for which the Gaussian model uses wind speed.

            2) Wind direction - representing the average wind direction at  10 m
above the ground  (and additional heights for elevated sources) during each  hour
calculated by carefully averaging samples of wind direction or by calculating
the resultant vector direction using unity as the wind speed for each sample.
A resultant vector direction does not represent the distribution of direction
samples which occurred during the hour.

            3) Temperature - representing the air temperature at the standard 2
m height above ground (and additional heights for elevated sources).

-------
                                                             Section No. 4.0.3
                                                             Revision No.     0
                                                             Date:   17 Sep 89
                                                             Page:    2  of  3


            4) Stability  class  -  representing  the  site of interest can be
estimated by a variety of schemes.   Turner  (1970)  describes a method based on
the observation of cloud  cover, ceiling height and wind speed along with a
known solar elevation angle.  Estimations may  also be based on the distribution
of wind direction (sigma  theta) or on  the vertical temperature gradient (delta
T).  Current research is  investigating whether or  not the Turner method
stability class can be estimated  with  measurements of solar radiation (daytime
sky cover substitute) and 2  to  10 m  delta temperature (nighttime sky cover
substitute) along with wind  speed, latitude and date. The method which will be
acceptable for the site of interest  is determined  by the regulatory authority.

            5) Mixing height -  may be  estimated by a method described by
Holzworth (1972).

            The PSD guideline provides accuracy and performance requirements
for wind speed, wind direction, temperature, vertical temperature difference,
and solar radiation along with  humidity, precipitation and visibility.

            Measurements  for PSD  permitting may, in some cases, be continued
after the new source begins  to  verify  the estimations made by modeling.
Continued monitoring requires the same QA/QC efforts as the permit phase
required.

            4.0.3.1.2  Other Programs

            Meteorological measurements may be made to augment air quality
measurements made to demonstrate  compliance with the National Ambient Air
Quality Standards (NAAQS) or to monitor trends.

       4.0.3.2  Research  Programs


       Meteorological data networks  may be  installed for special model
validation studies.  The  same kind of  QA/QC efforts are necessary for these
programs but they are usually applied  on a  shorter time scale since the
programs are relatively short in  duration and  the  need for documented accuracy
could not be greater.

       Data representativeness  is a  critical question as the terrain increases
in complexity.  Research  looks  into  the number and location of measurement
sites and the applicability  of  certain types of  instruments to characterize the
coaplex  (turbulent or stratified) air  flow  systems.  Different instruments,
such as Doppler sodars for remote vertical  sounding, sonic anemometers for
saall eddy size sensitivity  and low  threshold  speeds, laser anemometers for
long path length integration, and even the  old standby bivane, are examined to
try and optimize the detection  of important aspects of flow measurement for
model inputs or verification.

    Meteorological data are  used  to  find correlations with aerometric
measurements in a continuing search  for better forecasting capability.

       4.0.3.3  Contingency  Programs

       Industry may choose to monitor  meteorological variables at
representative sites on their property to document the local air flow

-------
                                                            Section No.  4.0.3
                                                            Revision No.     0
                                                            Date:    17  Sep 89
                                                            Page:     3   of  3


coalitions in case excessive concentrations are  observed which might
erroneously be attributed to their source.   While such  programs may not  fall
under any regulatory requirements, the use of  the data  for  its contingency
purpose  requires documentation sufficient to verify the data accuracy.

      Other programs may be exploratory to see  how on-site data compare with
public data from other sites (airports or state  or local agency stations).
Such questions of representativeness cannot be convincingly answered if the
oo-site  data does not come from suitably sensitive instruments, properly
calibrated and maintained and subject to QA/QC effort designed to document data
validity.

       It Is possible to select,  install, operate and document on-slte
measurement systems to meet PSD requirements.  Public data from airports may
differ from valid on-site data for three reasons.  Representativeness deals
witi different meso-scale structures in the surface layer  flow driven, in part,
by the larger synoptic flow.  It  is common to find that airport measurements do
not represent other sites Just a  few miles away because the flow is different.
It is also common to find airport data to be different  from on-slte data
because the airport data is essentially an instantaneous sample (a one minute
average) taken within ten minutes of the end of the hour while the on-site data
for the same hour Includes samples from the entire hour.   Finally,-the airport
instruments are  selected to serve aviation where  low wind speeds are  of no
importance.  Airport instruments  do not meet PSD  requirements.

-------
                                                             Section No. 4.0.4
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    1  of  7

4.0.4  MEASUREMENT REQUIREMENTS

       4.0.4.1  Measurement System

                4.0.4.1.1  Sampling

                The usual period of time assigned a data value is an hour.  It
is getting common to use a shorter intermediate period in the process of
generating the hour value.  Fifteen minutes is the recommended intermediate
period.   The fifteen minute values are usually calculated from samples taken
during the period.  The number of samples is related to the accuracy with which
the samples represent the true value for the period.   It has been found that
when the mean Is estimated by 60 samples, the sampling error Is 5 to 10
percent.  Also, when the standard deviation is estimated by 360 samples, the
sampling error Is also 5 to 10 percent.  For this reason,  the required number
of samples for sigma theta, the standard deviation of the wind direction is
equal or greater than 360 (EPA, 1987b).


                If a fifteen minute period is used and if 360 samples are
required within that period, a simple calculation shows the maximum time
between samples is 2.5 seconds.  How a sample is taken and what it represents
is also a consideration.  If a continuous output voltage is available, as with
wind direction, a sample of the voltage can be taken at any time.  If the wind
speed is calculated by measuring rate of rotation by counting pulses during a
fixed time, as is common for systems with the sensor directly connected to a
data logger (without a dedicated signal conditioner), the "sample" is really
the average for the fixed time.  If samples are taken once a second and the
aneaometer provides three pulses per revolution and the anemometer turns one
revolution for every 0.3 meters of air that goes through it, each pulse will
represent 0.9 m/s.  If samples are taken every 2 seconds,  the resolution of the
wind speed sample becomes 0.45 m/s.  A 15 minute period at 2 second sampling
will have 450 samples.  The average wind speed will be accurate with a
resolution of better than 0.1 m/s.   The variance of the wind speed samples may
be influenced by the 0.45 m/sresolution of the sample.

                Quality assurance considerations should include the
determination and documentation of the sampling procedures used in generating
the reported hourly data values.

                4.0.4.1.2  On-Llne Processing

                There are two on-line processing programs commonly used in air
quality meteorology.   One is the program used to combine wind speed and
direction samples for an hour.   The other is the program used to calculate or
estimate sigma theta.

                The QA role is to determine what these programs do and judge
trie suitability of the programs for the measurement application.   The field of
software QA for meteorology Is in it infancy and methods are not standardized
as yet.

                4.0.4.1.3  Data Handling

                There is a need to provide data in certain formats for some
applications.   If the data are machine processable In the  final measurement

-------
                                                             Section No.  4.0.4
                                                             Revision No.     0
                                                             Date:   17 Sep 89
                                                             Page:    2  of  7


step,  any reconfiguration  required will be handled by a program which can be
subject to software QA.

                If any hand  entry work is required, a data handling QC step is
required to be sure that errors  of transcription do not enter the data base.

       4. 0.4.2  Documentation

       If there is a requirement to  show evidence of data validity, the process
of documenting the various QA, QC, and operational activities is  important.
The added time such documentation takes is usually proportional to the degree
of preparation and training  which has been applied.

                4.0.4.2.1  Station Log

                The station  log  is the journal of all happenings  at the
aeasurement site.  These  Include visits where no problems are found, scheduled
calibration visits and findings, unscheduled maintenance tests and repairs, and
audits.  It is a truism that there are never enough field notes to reconstruct
with certainty what happened in  the  past.  Planning for the day when such a
reconstruction may be necessary  can  save a long period of data from being
discarded because of inadequate  documentation.

                4.0.4.2.2  Reports
                Any activity effecting the measurement system should be
reported.  This procedure  allows responsible individuals to follow these
activities without visiting  the  measurement site or witnessing calibrations and
audits.  It also provides  input  to a file of activities related to the system.
Reports should Include calibrations,  audits, discrepancies found  and corrected,
modifications or upgrades  and the like.  Reports do not need to be exhaustive
or glossy but they do need to be as  factual and succinct.


       4.O.4.3  Siting and Mounting

                4.0.4.3.1   Introduction
                Although  good instrumentation is a necessity, proper site
selection is critical to  obtain  good meteorological data.  It is, from an
absolute error point of view, much more important  than proper placement of  any
o-.ber kind of^air monitoring equipment.  Poor placement can and has caused
errors of 180° in wind direction, and can cause major errors in any other
meteorological variable,  including wind speed, temperature, humidity, and solar
radiation.

                The purpose  of this  section is to offer guidance  in assessing
tie suitability of meteorological monitoring sites.  The guidance given is
based principally on standards set by the World Meteorological Organization
(•MO, 1971). the Federal  Meteorological Handbook No. 1 (NWS, 1979) and the
Teanessee Valley Authority (TVA, 1977).  For an understanding of  flow around
obstacles and their potential bias  to wind data, see Hosker  (1984).

                Proper  siting is part of  the total quality control program.  Of
c=urse. as  in many other  monitoring  activities, the ideal may not be attainable
ard. in many urban areas  where air quality studies are traditionally done,  it
will be  impossible to find sites that meet all of  the siting criteria.  In

-------
                                                             Section No. 4.0.4
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    3  of  7

those cases, compromises must be made.  The important thing to realize is that
the data will be compromised, but not necessarily in a random way.  It is
incumbent upon the agency gathering the data to describe carefully the
deficiencies in the site and. if possible, quantify or at least evaluate the
probable consequences to the data.

                4.0.4.3.2  Instrument Siting
                The primary objective of instrument siting is to place the
instrument in a location where it can make precise measurements that are
representative of the general state of the atmosphere in that area, consistent
with the objectives of the data collection program.  Because most atmospheric
properties change dramatically with height and surroundings, certain somewhat
arbitrary conventions Bust be observed so that measurements can be compared.
In this section, conventions published by the World Meteorological Organization
(VMD, 1971) have been adopted wherever possible.  Secondary considerations such
as accessibility and security must be taken into account, but should not be
allowed to compromise data quality.
                     4.0.4.3.2.1  Wind Speed and Direction
                     "The standard exposure of wind Instruments over level,
open terrain  is 10 m above the ground" (WHO, 1971), however optimum measurement
height may vary according to data needs.  Open terrain is defined as an area
where the horizontal distance between the instrument and any obstruction is at
least ten (10) times the height of that obstruction.  An obstruction may be
man-nade (such as a building} or natural (such as a tree)(Figure 4.0.3.1).
                                 10T
Figure 4.0.4.1
Siting wind instruments; a 10 m tower located at  .'east
10 times the height of obstructions away from those
obstructions (not to scale).
                     The wind instrument should be securely mounted on a mast
that will not twist, rotate, or sway.   If It Is necessary to mount the wind
instrument on a roof of a building, it  should be mounted high enough to be out
of the area in which the air flow is disturbed by the building.  This is
us'.ially 1.5 times the height of the building above the roof so that it is out

-------
                                                             Section No. 4.0.4
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    4  of  7

of the wake  of the obstruction.  This is not a good practice,  however, and
should only  be resorted to when absolutely necessary.   Sensor  height and its
height above the obstructions, as well as the character of  nearby obstructions.
should be documented.

                      4.0.4.3.2.2  Temperature and Humidity

                      Temperature and humidity sensors  should be mounted over a
plot of open level ground at least 9 meters in diameter.  The  ground surface
should be covered with non-irrigated or unwatered short grass  or,  in areas
where grass  does not grow, natural earth.  The surface must not be  concrete or
asphalt or oil soaked.  The standard height for cllmatological purposes is 1.25
to 2 m, but  different heights may frequently be required in air quality
studies.

                      The sensors should not be closer  to obstructions such as
trees and/or buildings than a distance equal to four times  their height.  They
should be at least 30 o from large paved areas and not close to steep slopes,
ridges, or hollows.  Areas of standing water should also be avoided.  Louvered
instrument shelters should be oriented with the door opening toward true north,
in the northern hemisphere.

                      4.0.4.3.2.3  Radiation

                      Solar and whole sky radiation measurements should be taken
in a  location free from any obstruction to the measurements.   This means there
should be nothing above the horizontal plane of the sensing element that would
cast a shadow on it.  Neither should the instrument be near light colored walls
or artificial sources of radiation.  Usually a tall platform or roof of a
building  is  the most suitable  location.

                      4.0.4.3.2.4  Precipitation

                      A rain gage should be mounted on level ground so  that the
mouth or  opening is horizontal.  The gage should be shielded from the wind but
not placed  in an area where there will be excessive turbulence caused by the
shield.   For example, a good  location would be an opening in an orchard or
grove of  trees where the wind  speed near the ground is reduced due to  the
canopy effect, but a location  that is mostly open except for one or two trees
would not be good because of  the strong eddies that could be set up by  the
trees.  This admittedly requires a good deal of subjective  Judgment but it
cannot be avoided.  Obstructions to the wind should not be  closer than  two to
four times  the obstruction height from the instrument.  In  open areas, a wind
shield such  as that used by the U.S. National Weather Service should be used.
The ground  surface around the  rain gage may be natural vegetation or gravel.
It should not be paved, as this may cause splashing into the gage.   The gage
should be mounted a minimum of 30 cm above the ground and should be high enough
so that  it will not be covered by snow.

                      4.0.4.3.2.5  Meteorological Towers

                       It is frequently necessary to measure  some meteorological
variables at more  than one height.  For continuous measurements or where  the
height requirement is not too  restrictive, towers may offer the most
advantageous measurement platform.

-------
                                                             Section No.  4.0.4
                                                             Revision No.     0
                                                             Date:    17 Sep 89
                                                             Page:     5  of  7
                     Towers should be located in an open level area (see Table
4.3.4.1)  representative of the area under study.   In terrain with significant
topographic features,  different levels of the tower may be under the influence
of different meteorological regimes at the same time.   Such conditions should
be well documented.

     Table 4.0.4.1   Limits on Terrain and Obstacles Near Towers
Distance
Slope
from tower (between)
(m)
0- 15
15- 30
30-100
100-300

Source: TVA.
(X)
± 2
± 3
± 7
± 11

1977
Max. obstruction or
vegitation height

0.3
0.5-1.0
3.0
10 x ht.

(m)

(most

must
than


veg.<0.3)

be less
distance
to obstruction
                     Towers should be of the open grid type of construction,
such as is typical of most television and radio broadcast towers.  Enclosed
towers, stacks, water storage tanks, grain elevators, cooling towers,  and
   'lar structures should not be used (Mollo-Christensen, 1979).  Towers must
    ugged enough so that they may be safely climbed to install and service the
instruments.  Folding or collapsible towers that make the instruments available
to be serviced or calibrated at the ground are desirable provided they are
sufficiently rigid to hold the instruments in the proper orientation and
attitude during normal weather conditions.

                     Wind instruments should be mounted above the top of the
trver or on booms projecting horizontally out from the tower.  If a boom is
used, it should support the sensor at a distance equal to twice the maximum
diaaeter or diagonal of the tower away from the nearest point on the tower.
The boom should project into the direction which provides the least distortion
fcr the most important wind direction.  For example, a boom mounted to the east
of the tower will provide least distortion for north or south winds.  One may
wish to consider having two sets of instruments at each level, located on
opposite sides of the tower.  A simple automatic switch can choose which set of
data to use (NASA, 1968).  Documentation of the tower should include the
orientation of the booms.

                     Temperature sensors must be mounted on booms to hold them
a«ay from the tower, but a boom length equal to the diameter of the tower is
sufficient.  Temperature sensors should have downward facing aspirated shields.
The booms must be strong enough so that they will not sway or vibrate
excessively in strong winds.  The best vertical location on the tower for the
sensors is at a point with a minimum number of diagonal cross members, and away
frrm major horizontal cross members.  Even with these precautions, data
   "Ined while the wind blows from the sector transected by the tower may not
    ree from error.
                     These instrument siting suggestions may seem to preclude
t-e use of many air monitoring sites that otherwise would be desirable, but

-------
                                                             Section No. 4.0.4
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    6  of  7


this need not  be  the  case.   In siting  air quality  monitors that are to be used
for long-term  trend analysis or large  geographic area coverage, it may be
perfectly acceptable  to have some or all of  the meteorological equipment at a
different location that better meets the large-scale requirements of the study.
As long as both sites are in the same  area of  interest and meet their
respective siting criteria,  this should present no problems.  When the air
quality data are  to be used for short-term diffusion model validation or
studies of short-term levels from specific sources, however, a meteorological
station should be located in the vicinity of the air quality sensor.

                4.0.4.3.3  Station Siting

                Besides care in selecting the local environment of a
meteorological sensor, it Is important that  care be taken  in selecting station
location with  respect to major man-made and  topographic  features  such as
cities, mountains,  large bodies of water, etc.  Meteorological variables are
obviously affected by the large-scale  surrounding  features.  The  effect of
cities has been studied extensively (Ito, 1972; Vukovich,  1971; U.S.PHS, 1961).
Documented effects  include a decrease  in an  average wind speed, decrease in
atmospheric  stability, increase in turbulence,  increase  in temperature, and
changes in precipitation patterns.  These changes  will obviously  have an effect
on the evaluation and interpretation of meteorological and air quality data
taken in an  urban area.

                Even more pronounced are  the effects of  large natural features
(Slade, 1963). Besides their obvious  effect on humidity,  oceans  and  large
bodies of water are  usually at a different  temperature than the nearby  land.
This generates the well known land- and sea-breezes which,  in many coastal
areas, dominate  the  wind patterns.  There  are also often simultaneous
differences  in cloud cover between the oceans and  nearby land surfaces.  This
difference  in thermal lag. Insolation, and  changes in surface roughness  and
vertical  temperature structure can have a  profound effect  on atmospheric
stability  (SethuRaman. 1974).

                 The effects of mountains and valleys on  meteorological
variables  and atmospheric dispersion continue to  be studied.  Two of  the more
interesting  recent  papers are by Kao  (1974)  and Hunt  (1978).  Well-known
effects include the channeling of flow up or down a valley,  the creation of
drainage  flow, the  establishment of lee-waves, and an  increase  in mechanically
generated  turbulence.  All of these effects and others can play a major role  in
the  transport and dispersion of pollutants.

                 The important point is that almost any physical object  has an
effect upon  atmospheric motion.   In fact,  it is probably impossible  to  find a
site that  is completely free from obstruction.  This being the  case,  it is  the
responsibility of the person choosing a monitoring site  to have  In  mind the
various forces at work and  to choose a site that will  be most  representative  of
the  area  of  interest.   If the area is a valley or a sea  coast,  then the
meteorological instruments  should be  in that valley or near the  coast;  not on  a
nearby hilltop or inland 30 km  at a more convenient airport site.  Of course
one  must  also keep in mind  the  vertical structure of the atmosphere.   Winds
measured  at  the bottom of a  100  m valley will not represent the  winds at the
top  of a  ZOO m stack whose  base  happens to be .'n that  valley.

-------
                                                             Section No. 4.0.4
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    7  of  7


                The choice of a station for meteorological data collection must
be made with a complete understanding of the large-scale geographic area, the
sources being investigated, and the potential uses of the data.  Then rational,
informed choices can be made.

                Once they are made, the site should be completely documented.
This should include both small- and large-scale site descriptions, local and
topographic maps (1:24,000 scale), photographs of the site (if possible), and a
written description of the area that is adequately represented by this site.
This last point is most important for it will allow a more rational
interpretation of the data.  It might state, for example, that a site
adequately represents a certain section of a particular valley, the suburban
part of a given city, or several rural counties.  Whatever it is, the nature of
the site should be clearly described in a way that will clear to those who will
be using the data in the future.

-------
                                                             Section No. 4.0.5
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    1  of  2
4.0.5     REFERENCES
EPA.  1976:   Quality Assurance Handbook for Air Pollution Measurement
    Systems, Vol.  I, Principles, EPA-600/9-76-005.   Office of Research and
    Development, Res.  Triangle Park, NC 27711.

EPA.  1977a:  Quality Assurance Handbook for Air Pollution Measurement
    Systems, Vol.  II,  Ambient Air Specific Methods,  EPA-600/4-77-027a.  Office
    of Research and Development, Res. Triangle Park, NC 27711.

EPA.  1977b:  Quality Assurance Handbook for Air Pollution Measurement
    Systems, Vol.  Ill, Stationary Source Specific Methods,
    EPA-600/4-77-027b.  Office of Research and Development,
    Res.  Triangle Park, NC 27711.

EPA.  1982:   APTI Course SI:  409, Basic Air Pollution Meteorology, Student
    Guidebook EPA 450/2-82-009.  Air Pollution Training Institute.
    Environmental Research Center, Research Triangle Park, NC 27711.

EPA.  1983:   Quality Assurance Handbook for Air Pollution Measurement
    Systems, Vol.  IV,  Meteorological Measurements,  EPA-600/4-82-060.  Office
    of Research and Development, Res. Triangle Park, NC 27711.

EPA.  1986:   Guideline on Air Quality Models (Revised).  OAQPS. U.S.
    Environmental Protection Agency, Research Triangle Park, NC.,
    EPA-450/2-78-027R (NTIS PB 288-783).

EPA.  1987a:  Ambient Monitoring Guidelines for Prevention of
    Significant Deterioration  (PSD), EPA-450/4-87-007, Office of Air
    Air Quality Planning and Standards, Res. Triangle Park, NC 27711

EPA.  1987b:  On-Site Meteorological Program Guidance for Regulatory
    Modeling Applications, EPA-450/4-87-013. Office of Air
    Quality Planning and Standards, Res. Triangle Park, NC 27711.

Holzworth,  G. C..  1972:  Mixing Heights, Wind Speeds, and Potential for
    Urban Air Pollution Throughout the Contiguous United States.  Office of
    Air Programs.  U.S. Department of Health, Education and  Welfare, Research
    Triangle Park, NC.  Publication Nol. AP-101.

Hosker, Jr.,R. P..1984:  Flow and Diffusion Near Obstacles, Chapter 7.
    Atmospheric Science and Power Production,   Darryl Randerson,
    Ed. DOE/TIC-27601, pp 241-326.

Hunt, J.  C. R., W. H.  Snyder and R. E. Lawson, Jr.,  1978:  Flow structure
    and turbulent diffusion around a three dimensional hill:  part 1.
    EPA-600/4-78-041.  U.S. Environmental Protection Agency.

I to,  Masashi et al., 1972:  An examination of local wind measurements in
    cities.  Annual Report of the Tokyo Metropolitan District Public Damage
    Research Institute,3(3). pp. 27-32  (APTIC 41974 TR 186-73).

-------
                                                             Section No. 4.0.5
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    2  of  2


Kao, S. K. , H.  N.  Lee  and K.  I.  Smidy,  1974:  A preliminary analysis of the
    effect of mountain-valley terrains  on turbulence and diffusion.
    Symposium on Atmospheric Diffusion  and Air Pollution, American
    Meteorological Society,  Santa Barbara, CA, pp. 59-63.

Mollo-Chrlstensen, E.,  1979:   Upwind  distortion due to probe support in
    boundary layer observations.   Journal of Applied Meteorology, 18(3),
    pp. 367-370.

NASA, 1968:  Meteorological  Measuring and Recording Equipment
    Description,  Calibration and Maintenance Procedures for NASA's 150
    Meter Meteorological  Tower Facility,  Kennedy Space Center, Florida,
    National Aeronautics  and Space Administration, GP-465.

NWS, 1977:  Federal Meteorological Handbook No. 1, Surface Observations.
    U.S. Department of Commerce,  Washington, D.C.

SethuRaman, S. ,  R. M.  Brown,  and J. Tichler, 1974:  Spectra of
    atmospheric turbulence over the sea during stably stratified
    conditions.   Symposium on Atmospheric Diffusion and Air Pollution,
    American Meteorological  Society.  Santa Barbara, CA, pp. 71-76.

Slade. D. H.,  1968:  Meteorology and  Atomic Energy.  U.S. Atomic Energy
    Commission.  TID-24190.

Turner, D. B. ,  1970:   Workbook of Atmospheric Dispersion Estimates,
    Revised.  Office of Air  Programs, U.S. Department of Health, Education
    and Welfare,  Research Triangle Park,  NC.  Publication No. AP-26.

TV A, 1977:  Criteria for  meteorological measurement site acceptance
    and/or preparation (unpublished).   Tennessee Valley Authority.
    U.S.Public  Health  Service, 1961:  Air Over Cities. A symposium,
    Cincinnati,  Ohio.

Vukovich, F. M. ,  1971:  Theoretical analysis of the effect of mean wind
    and stability on a heat  Island circulation characteristic of an urban
    complex. Monthly Weather Review,  99(12), pp. 919-926.

WMO, 1971:  Guide to meteorological instrument and observing practices.
    World Meteorological  Organization No.  8TP3, 4th edition, Geneva,
    Switzerland.

-------
                                                            Section No.  4.1.0
                                                            Revision No.     0
                                                            Date:    17 Sep  89
                                                            Page:     1   of   3

                                  Section 4.1
                    GENERAL ASPECTS OF QUALITY ASSURANCE FOR
                          METEOROLOGICAL MEASUREMENTS
                                    OUTLINE

Section                                                    Pages Rev. Date

4.1.0  OUTLINE AND SUMMARY                                  3     0  9/89

4.1.1  PLANNING FOR A QUALITY ASSURANCE PROGRAM             3     0  9/89

       4.1.1.1  PROJECT DESCRIPTION
       4.1.1.2  PROJECT ORGANIZATION
        .1.1.3  QA OBJECTIVE
        .1.1.4  CALIBRATION METHOD AND FREQUENCY
        .1.1.5  DATA FLOW ANALYSIS
        .1.1.6  VALIDATION AND REPORTING METHODS
        .1.1.7  AUDITS - PERFORMANCE AND SYSTEM TYPES
        .1.1.8  PREVENTIVE MAINTENANCE
        .1.1.9  QA PROCEDURES
        .1.1.10 CORRECTIVE ACTION AND REPORTS

4.1.2  ORGANIZATION OF AUTHORITY AND RESPONSIBILITY         1      0  9/89

4.1.3  QALITY CONTROL GUIDELINES                            2     0  9/89

4.1.4  TRACEABILITY PROTOCOL                                4     0  9/89

       4.1.4.1  VOLTAGE
       4.1.4.2  WIND SPEED
       4.1.4.3  WIND DIRECTION
       4.1.4.4  TEMPERATURE AND TEMPERATURE GRADIENTS
       4.1.4.5  SOLAR RADIATION
       4.1.4.6  ATMOSPHERIC WATER VAPOR
       4.1.4.7  PRECIPITATION
       4.1.4.8  ATMOSPHERIC PRESSURE

4.1.5  ESTIMATING PRECISION AND ACCURACY                    6      0  9/89

       4.1.5.1  DEFINITIONS
       4.1.5.2  COLLOCATED TRANSFER STANDARDS
       4.1.5.3  OTHER CONSIDERATIONS

4.1.6  SYSTEM AUDITS                                        2      0  9/89

4.1.7  PERFORMANCE AUDITS                                   1      0  9/89

4.1.8  DATA VALIDATION PROTOCOL                             7      0  9/89

       4.1.8.1  STRIP CHARTS
       4.1.8.2  METHODS
       4.1.8.3  THE EMSL/RTP SYSTEM

-------
                                                            Section No.  4.1.0
                                                            Revision No.     0
                                                            Date:   17 Sep 89
                                                            Page:    2  of  3

4.1.9  QA REPORTS AND CORRECTIVE ACTION                     2    0   9/89

       4.1.9.1  OPERATIONS LOG AND MAINTENANCE REPORTS
       4.1.9.2  CALIBRATION REPORTS
       4.1.9.3  AUDIT REPORTS
       4.1.9.4  REPORTS TO MANAGEMENT
       4.1.9.5  DISCREPANCY REPORTS

4.1.10 REFERENCES                                           2    0   9/89

-------
                                                             Section No. 4.1.0
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    3  of  3

                      GENERAL ASPECTS OF QUALITY ASSURANCE
                        FOR METEOROLOGICAL MEASUREMENTS
                                    SUMMARY

    Quality assurance  (QA) for meteorological measurements is a relatively new
field.  There are generally two reasons for recording meteorological data.  One
is to learn what the atmosphere is doing, particularly the lower part of the
boundary layer.  The other is to document what  the atmosphere is doing.  It is
necessary to find the  relevant facts and understand them to learn something .
To document what has been learned may require considerably more data and will
require some assurance the data are correct.  The organizations that need valid
data are the ones which collect it, and they will write their own procedures.
This is how the National Weather Service has handled data collection for
synoptic and climatological applications.  When  third party requirements with
the force of law began to need meteorological data for transport and diffusion
modeling and safety analysis, the need for QA was established.

    As with most specialties, HA in general has  acquired its own language and
infrastructure.  In this handbook, the goal is  to avoid structure which has no
specific value to meteorological data validity.  On the other hand, the goal is
to provide clear definitions, methods and examples which will help produce and
verify valid meteorological data.  Some of the popular sayings or phrases make
valid points.  The book "QUALITY IS FREE" (Crabby. 1979) promotes the idea that
it is really cheaper to do it right the first time.  This concept is easy to
demonstrate with manufactured products where bad products will either cost more
through warranty repairs or lost sales and bad will.  If the "product" is data
or services producing  data, an awareness of the  ultimate cost of the loss of
data is important.  If no one ever looks at or uses the data, it is a waste
of money to buy. install and operate instruments and recording systems.  Even a
facade of data is not  worth the money.  If there is a reason for meteorological
measurement, that reason will provide the basis  for estimating the economic
down-side for producing unacceptable data.

    Of course data judged "unacceptable" must have been rejected by someone for
some reason.  The reason for needing the data in the first place will provide
the basis for the economic price which must be 'paid for either not "doing it
right the first time"  or for fighting the rejection because it was done right.
Usually the reason for measuring meteorological variables is a government
regulation requirement or a need to combat potential claims of injury.  It can
be argued that the fagade of data coming from the instruments on the tower will
satisfy the government regulation requirement.   It has been argued that an
extra nickel spent measuring meteorological variables (or air quality, for that
matter) is a nickel lost to the bottom line profit of a manufacturing plant.

    The assumption upon which this handbook rests is that the pertinent
government regulations and guidance documents will clearly define what valid
data are and how validity is proven.   And further, that those people
responsible for accepting or using the data will require that the data be valid
or rejected.  The expertise necessary for this determination will be found in
this handbook and its  references.   This should equally help those who must
collect valid data and those who must accept the claim of validity.   This
assumption clearly moves the purpose for collecting data away from learning and
into the category of documenting.   Documentation needs QA.

-------
                                                             Section No. 4.1.1
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    1  of  3
4.1.1   PLANNING FOR A QUALITY ASSURANCE PROGRAM
    A formal quality assurance program should be designed into the monitoring
program so that provisions can be made in the measurement system design for
necessary quality control checks and for better monitoring of system
operations.   If these activities are planned and provided for by incorporation
of special readouts, calibration equipment,  spares,  and procedures for their
use. then the system Is more likely to perform in a satisfactory manner and
deliver valid data.

    The formal plans for quality assurance are presented in a document called a
QA Plan (Lockhart. 1985c and EPA, 1987b).  This plan lists all the
quality-related procedures and the frequency of their use to document the
operation of the instrument system.   The QA Plan contains information under
different headings to organize all the various activities in a logical sequence
and to avoid overlooking an important step.   The specifics of each plan must
relate to the needs of the program,  but the general content elements are the
saae.
       4.1.1.1 Project Description

       This introduction establishes why the documentation of meteorological
data monitoring is needed and why it is important to the organization that
valid data are collected.  It also describes how the data will be used which
establishes the criteria for judging the representativeness of the data.

       4.1.1.2 Project Organization

       The literature of QA abounds with examples of the importance of well
defined organizational structure starting with the organization policy on
quality, endorsed in writing by top management.  This provides the authority to
"do it right the first time."  If the organization has no policy on quality and
if someone at the operations level is given the QA responsibility without
sufficient authority (often the case), the effort may become just another
secondary task which must be done.  This is an Invitation to a facade system.
An organization will seldom build a new plant without the expertise of
architects and engineers.  Meteorological data systems are often assembled from
parts picked from catalogs by experts in other fields who do not understand the
routine operating requirements for collecting valid data.  A valid QA Plan is a
structure to encourage and guide organizations toward a successful collection
of seeded data.

       4.1.1.3 QA Objective

       This section is the real QA plan.  The first two sections described the
project for which the data will be used and the organization of those who will
par-iclpate In the data collection.   This section contains the details of how
the QA program will monitor the collection process with the purpose of
doc-jnenting and defending claims of data validity.

       4.1.1.4 Calibration Method and Frequency

       Calibrations are tests, and adjustments if necessary, to relate the
instrument system to truth or validity.  The evidence of this activity, the
documentation, is the foundation upon which the judgment of data validity must

-------
                                                             Section No.  4.1.1
                                                             Revision No.     0
                                                             Date:   17 Sep 89
                                                             Page:    2  of  3


rest.   This section defines in advance how the  calibrations will be done,  how
often and by whom.

       4.1.1.5 Data Flow Analysis

       This section starts with samples of atmospheric conditions, a rate of
rotation of an anemometer representing wind speed for example, and describes
how the samples  are combined Into reported values.   The  section describes how
these values, perhaps  hourly averages, are Inspected and Judged to be
acceptable or not.  Finally, after validity has been established, the data are
archived in some way to become available for use as the  project requires.   An
experienced meteorologist reading this section  will know what  the data mean and
what data quality control has been applied.

       4.1.1.6 Validation and Reporting Methods

       Section 4.1.1.4 provides points in time  at which  the  instrument
performance is known.   This section describes what criteria  are used for any
automatic data Inspection programs applied between calibrations and how the
results of such  programs are implemented and reported.   If comparisons are made
to other similar measurements, such as a wind speed at a different location or
a different time, this section will document the methodology applied.

       4.1.1.7 Audits  - Performance and System  Types

       Audits may be  required or chosen to add  to the documentation some
independent evidence of the performance of the  meteorological  instruments
and/or the performance of those who are responsible for  implementation of  the
QA Plan.  This section defines how often performance audits  are used to
challenge  the measurement instrument system and how often system  audits are
used  to challenge the  implementation of the QA  Plan or program.   Also defined
is the type of auditor to be used.  Internal auditors are members of the
organization who are  independent from those responsible  for  collecting and
handling  the data.  External auditors are usually outside contractors.  In
either case,  the auditor must be experienced in the field of meteorology and
must  be provided support from the operating organization. Auditing should be
the most positive learning  experience for operators and  a contributor to data
validity.

       4.1.1.8 Preventive Maintenance

       Some  instruments require routine service to assure valid data.  Solar
radiation  sensors have glass covers which need to be properly  cleaned on some
schedule  (daily  or weekly), depending on its location.   Tipping bucket
raingages  need  to be checked periodically for spiders  or other insects which
might take  up residence in  the bucket mechanism.  Anemometers  and wind vanes
have  bearings which will need service  (usually replacement)  on some time scale
 (quarterly or annually), depending on  the environment.   Some dew  point sensors
require  coating periodically.  All such predictable service  points should  be
recognized  and a preventive maintenance plan described for each of  them in this
section.

-------
                                                             Section No.  4.1.1
                                                             Revision No.    0
                                                             Date:    17 Sep 89
                                                             Page:     3  of  3
       4.1.1.9 QA Procedures
       Section 4.1.1.2 describes the QA objectives.   This section contains the
details of how these objectives will be met.   A written procedure will both
document how the QA task will be addressed and guide a QA person through the
process.   Procedures are a mechanism for establishing technically correct
methodology which can be followed by people without  the technical background or
experience to write the procedure.   While it is not  practical to use experts to
perform routine tasks, it is necessary to have the expert guidance to follow.
Procedures fill this need.  Procedures should be controlled to the degree that
they cannot be changed without written approval of an expert.  A system audit
checks to see if procedures are being followed as they are written.

       4.1.1.10 Corrective Action and Reports

       Documentation is the main goal of a QA Plan.   General procedures will
require noting in the Site Log any activity relating to the meteorological
system.   Top management, having established the policy and granted the
authority to "do it right the first time," needs to  be aware of the QA
activities required by the QA Plan.  If a procedure  or inspection uncovers a
discrepancy with respect to the clearly written system specifications, a
discrepancy report becomes the message to initiate corrective action.  Top
management needs to see these reports along with the corrective action
statements (usually a part of the discrepancy report form) to know that the
system is in control.  Too often, problems must be visible to top management if
corrective action to the system is going to be initiated.  Audit reports and
other performance reports are circulated and followed up by procedures
described in this section of the QA Plan.

-------
                                                             Section No.  4.1.2
                                                             Revision No.    0
                                                             Date:    17 Sep 89
                                                             Page:     1  of  1
4.1.2  ORGANIZATION OF AUTHORITY AND RESPONSIBILITY
    Quality does not mean the best money can buy.   It means that the customer
gets what he needs and expects, no more and no less.   The way to assure a
quality product or service is to first set a top management policy in writing
stating that goal or commitment.  The implementation of such a policy requires
a person who can avoid the conflict of interest of providing the product or
service and Judging its quality before it is delivered.  This usually means a
Quality Assurance Manager.  Once the policy is set and the QA Manager is chosen
the procedures by which quality is assured can be written, usually a QA manual.
The manual might proclaim that every project with a deliverable product or
service will have a QA Plan.

    The authority to establish this kind of organizational structure must be
top management.  During the establishment period,  top management must
participate and approve the quality organizational structure and procedures.
Once established, top management can delegate authority to the QA Manager for
operating the department.  A routine feedback from the QA Manager to top
management is necessary to preserve the control of delegated authority, see
that it is being used effectively and demonstrate to the rest of the
organization a level of importance placed on the quality policy.

    Once the organization is in place, and QA Plans are required, the Job of
collecting wind data for a PSD  (Prevention of Significant Deterioration)
application starts with a QA Plan.  What are the measurement requirements?  How
will the data be handled?  How will the instruments be calibrated and serviced?
What procedures will be used?  What outside authority will assure management
that the QA Plan is adequate?  Once all the steps required to gather a year of
valid wind data have been defined, the QA people will monitor the process, help
train inexperienced operators, and build a documentation base to support the
claim of validity for the data once the year is over.  This should include 'some
outside auditing to add to the documentation an impartial expert opinion of
satisfactory performance, or bring to the attention of the QA people and their
management any problems which might have been overlooked.  The best time for an
independent audit is at the beginning of the data-taking period when the loss
of valuable time can be minimized.

    The principles of this approach to quality are sound and irrefutable.  The
cost is less than any other approach.  All that is required is to figure out,
in sufficient detail, what and how the Job is to be done before it begins and
to specify how the Job will be monitored to assure satisfactory completion.
Anything less is a gamble which may or may not pay off.  Host organizations do
not like to gamble, but many do.

-------
                                                             Section No.  4.1.3
                                                             Revision No.     0
                                                             Date:    17 Sep 89
                                                             Page:     1  of  2
4.1.3  QUALITY CONTROL GUIDELINES
    Quality control Is a process which operates  In  parallel  with the production
of a product or a service.  There is a gigantic  body  of  literature on the
subject.   Some examples of books are Juran (1979),  Feigenbaum (1961) and Grant
and Leavenworth (1974).  A technical organization,  the American Society For
Quality Control (ASQC) exists for the promotion  of  quality systems.  ASQC has
committees of volunteers to establish guidelines and  standards for the quality
profession.  One such committee is the American  National Standards Institute
Accredited Standards Committee (ANSI ASC) Z-l  Committee  on Quality Assurance.
A product of this committee is a series of standards  (ANSI/ASQC 1987a-e) in
which is stated the need to qualify quality control with an  adjective
describing what is to be controlled.  This need  is  nowhere greater than in
meteorological monitoring.

    In the meteorological literature there are recent papers (Wade, 1987 and
Lockhart, 1988) in which the use of quality control is discussed.   There is a
difficulty with the language used to describe  a  task  and the perception of
control as related to those working the task.  Assume the task is the
accumulation of a one year data set of wind and  temperature  measurements.  The
QA Plan describes the goals and the specifications  to which  the instruments
must conform.  Purchasing has bought the instruments  with a  purchase order
requiring conformance to performance specifications and  describing how the
conformance will be tested.  A suitable site has been found  and a consensus has
been reached as to the representativeness of the site.   Receiving Inspectors
accepted the instruments and operators have installed them.   The QA Plan called
for an Independent performance audit at the beginning of the data year which
the system passed.  The QA Plan calls for an inspection  of the data on a weekly
basis by the meteorologist or environmental scientist who will be working with
the data.  The QA Plan provides a procedure with which  the data inspector can
communicate in writing with the operators to report questionable data and
receive an answer of special instrument checks.   The  QA  Plan requires operators
to calibrate the instruments on a six month frequency or when problems are
found.  The QA Plan specifies how the calibrations  are  to be done and to whom
the reports will be routed.  At the end of the year the  data are summarized and
made ready for use in diffusion models.  Where is control and where Is quality
control?

    The whole program is controlled by the top management through the QA Plan.
How well the various parts of the organization carry out their responsibilities
is checked on by the QA people or person.  If  the receiving  inspectors balk at
performing their service because they are too  busy, top  management participates
in the decision to either modify the QA Plan and Policy  on Quality or find a
way to accomplish the receiving inspection. Management  may  choose to gamble
that the instruments are all right and any problems will be  uncovered during
the installation and audit.  This is management's prerogative and this gamble
is a pretty safe one.  Without the QA Plan the manager of receiving may make
the decision without the benefit of knowing what the  stakes  of the gamble might
be.  While the whole project is controlled by  the QA Plan, there is one data
quality control function performed by the meteorologist/environmental
scientist; on a weekly basis the data are examined  and accepted or rejected.
If problems or questions arise, discrepancy reports will be  initiated which
operate in accordance with the QA Plan.  The inspector  finds that the wind
direction looks too steady and writes it up.  The operator goes to the site

-------
                                                             Section No. 4.1.3
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    2  of  2

and, finding  that  the  vane has been removed by  a  predatory bird,  installs a
spare, notes  the action in the site log and on  the discrepancy  report.
recalibrates  if necessary, and reports back to  the inspector closing the loop.
This is a true quality control activity.   The quality of  the data is being
controlled by the  experienced Judgment of the data inspector who  looks back in
the data to find when  the vane disappeared,  and deletes or flags  the data as
Missing.

    Another quality  control function is the periodic calibration.  This is an
instrument quality control process where adjustments are  made as  necessary to
keep the instruments "in control."  If more frequent calibrations are used or
if the program goes  on for several years, a standard control chart may be used
to visually track  the  "In control" status of the  instruments.   The data quality
control Inspector  must see the calibration reports and contribute to the
decision about what  if anything to do to the data as a result of  calibration
findings.  Doing anything to the data requires  very careful consideration and
thorough documentation.

    When the  organization policy is to achieve  the level  of quality the
"customer" expects,  the whole organization effectively becomes  a  part of the QA
department.   Various techniques, such a quality circles,  may be used to
maintain a high level  of quality through broad  participation and  training.
These techniques also  underscore the management's dedication to the quality
policy.  It is only  when other criteria,  such as  departmental profit goals,
enjoy a higher priority than does quality that  an Independent "watch dog"
organization  is required to achieve a published quality policy.   The price for
quality is inversely proportional to its place  on the priority  scale.  When
quality is first on  the list, the most efficient  and least expensive process
can be found  for its achievement.

-------
                                                             Section No. 4.1.4
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    1  of  4
4.1.4  THACEABILITY PROTOCOL
    There is a general practice in QA/QC to use a hierarchy of standards
resting on international standards or those maintained by the National Bureau
of Standards (NBS).  This results in claims of calibrations that are "traceable
to NBS."  While it is difficult to define traceability in quantitative terms,
there is value in using an authority against which other instruments can be
compared.  This section will discuss how this hierarchy relates to
meteorological measurements.

    Most meteorological measurements contain some sensing element which reacts
to the variable of interest and the usual transducer outputs of voltage,
current or frequency..  In terms of accuracy, the response of the sensing
element is the most  important and the most difficult to define (see sub-section
2, Specifications, for each variable).  The measurement of the various
components of the electrical output, including digital code, is straight
forward and subject  to normal methods of calibration and certification.
Protocols for "traceability to NBS" for voltage will be discussed first.

       4.1.4.1  Voltage

       Regular calibration labs maintain transfer standards which are sent  to
NBS for calibration.  These in turn are used to calibrate the lab's voltage
sources which in turn are used to calibrate a subject volt meter.  This process
has gone on for years and is called "traceable to NBS."  Calibration labs check
a volt meter, adjust It  if necessary, and affix a calibration sticker
certifying the meter to be In calibration at the date tested and recommending
re-certification at  a future date  (six months to a year).  This traditional
process is entirely  acceptable.

       Manufacturers of volt meters also have transfer standards which they use
to calibrate and certify  their products.  Modern digital volt meters or
volt-ohm meters  (DVOM) are very stable in calibration, particularly those of
high quality.  Another method for achieving "traceability to NBS" Involves  the
comparison of DVOMs.  If  one DVOM  is certified as accurate, either by a
calibration lab or by the manufacturer with a transfer standard, and another
DVOM is placed  in parallel across a voltage source, the uncertified one can be
certified by that comparison.  This process is valuable to use in performance
auditing in order  to fix  the accuracy with which the operator calibrates the
signal conditioners.  The process yields a relative accuracy if neither DVOM is
certified, but  it becomes absolute as comparisons are made with certified
DVOMs.

       What is an acceptable error in voltage measurement for meteorological
purposes?  Assume a  measurement system with a full scale output of 1 volt for
wind speed, wind direction and temperature difference.  Assume the DVOMs are on
a range displaying millivolts where the  1 volt full scale looks like 1.000.
How important  is  it  if  the two DVOMs disagree by as much as 0.002 volts?  If
the range for wind speed  is 0.2 to 50 m/s, and the accuracy requirement is  0.2
m/s, what is that accuracy requirement expressed in volts?

                50.0  m/s  • 1.000 V
                0.2  m/s  = 0.2/50 x 1.000 = 0.004 V

-------
                                                             Section No. 4.1.4
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    2  of  4

The disagreement between the DVOMs is  equivalent to half the accuracy
requirement.  For all  practical  purposes a disagreement of this size is not
important, but  an auditor would  like more information.  Is it a bias or a
random difference?

       A good DVOM.  like the Fluke 8060A, specifies Its accuracy on the 2.0000
V range as ±  (0.04X  of reading + 2 digits).  On the 20.000 V range the accuracy
is specified as ± (0.05% of reading +  2 digits).   If the two DVOMs were on the
same output of  0.1000  V (5 m/s for the wind speed  example), and if they were on
a range equivalent to  the 2 V range stated above,  they should each read 0.1000
±(0.00004 + 0.0002)  or between 0.09976 V and 0.10024 V.  Truncation of the
measurement to  fit the display would cause the meters to read between 0.0997
and 0.1002 which would be interpreted  as 4..98 and  5.01 m/s.  If the difference
were as much as 2 mV (0.1 m/s),  it would indicate  a bias (calibration) error in
one or both of  the DVOMs.  If the DVOMs were each  on a 20 V range, they should
read 0.10O ±  (0.00005  + 0.002) or between 0.09795  V and 0.10205 V.  Truncation
would force the meters to read between 0.097 and 0.102 which would be
interpreted as  4.85  and 5.10 m/s.  A difference of 2 mV (0.1 m/s) in the meter
readings could  be either a bias  error  or a random  error from the 2-digit
uncertainty.  Switching both DVOMs to  the 2 V range would resolve the question.

       Table  4.1.4.1  summarizes the accuracy of the conversion of the
transducer output to voltage output in units of voltage and units of
meteorology for the  1  volt range example for wind  speed, direct Ion and
temperature difference.

                 Table  4.1.4.1 - Voltage vs.  Met.  Unit Accuracy
Variable
Range
Volts
Wind speed
Direction
A T
0.
0.
0.
000 -
000 -
000 -
1
1
1
.000
.000
.000
Met.
0.0
360
-5.0
Units
- 50.0
- 360
- 15.0
Accuracy (0.1% FS)
Volts
m/s
deg.
C
0.
0.
0.
001
001
001
Met. Units
0.05
0.36
0.02
m/s
deg.
C
       4.1.4.2  Wind Speed
       Traceability to NBS has some meaning in  the  measurement  of  wind speed.
The National Bureau of Standards Fluid Mechanics Section operates  a pair of
calibration wind tunnels at their facility in Gaithersburg, MD  (Washington,
D.C.).  One can arrange to send an anemometer to NBS for calibration.  A report
will result which describes the output of the sensor or system  (rate of
rotation or volts) at a series of wind speeds.  NBS states  the  accuracy of the
wind speed  they use to be 0.1 mph.   How the user Implements the test report  is
a different story (see 4.2.12).  If the user is a manufacturer,  the test report
will probably be smoothed by some least square  method which predicts speed from
rate of rotation.   The speed predicted by the rate  of rotation  of  the
anemometer  calibrated by NBS will then be transferred to another anemometer  by
collocating them in another wind tunnel or by calibrating the wind tunnel as an
intermediate standard.  If new anemometers agree with this  transfer of the
performance of the "standard" anemometer to within  some margin  of  error, the
calibration of the new anemometer is said to be "traceable  to NBS."

-------
                                                             Section No. 4.1.4
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    3  of  4


    There Is no standard wind.   NBS uses a structure designed to smoothly
control  the air being driven by a propeller-motor assembly.   How smoothly and
uniformly the air flows through the test section Is determined by testing.
The wind speed at some point In the test section is calculated from the
measured pressure difference between the pltot tube and the static pressure,
correcting for air density.   The pressure difference is measured with a
manometer.   Anyone can build a wind tunnel and measure its performance as
accurately as can NBS.  "Traceable to NBS" provides a relative standard of
comparison with absolute errors which are small compared to the needs of the
scientific and industrial users.

      4.1.4.3 Wind Direction

      "Traceable to NBS" has no meaning as it relates to wind direction
(see 4.2.2.2).

      4.1.4.4 Temperatureand Temperature Gradients

      There is a hierarchy for temperature much the same as voltage.
Calibration labs and manufacturers maintain sensors with calibrations run by
NBS.  A  user can send an electrical transducer, which has a unique relationship
between  resistance and temperature, to a calibration lab and get a report on
that relationship as determined by the lab's transfer standard.  This
calibration is called "traceable to NBS" because the transfer standard was
calibrated there.  Some concern about how the subject transducer and the
transfer standard are exposed to the "same" temperature is warranted.  The test
nethod and test facilities are not usually certified by NBS and so the
calibration may not deserve the Inferred NBS authority.

      Differential temperature is nothing more than two or more temperature
measurements taken at different points.  The important calibration for this
variable is one which compares one instrument to the other,  a
relatlvecallbration.  Traceabllity Is not relevant to relative calibrations.

      4.1.4.5  Solar Radiation
                *
      Traceabllity to an absolute measurement of solar radiation is achieved
by collocated comparisons with secondary standards at organizations such as the
Desert Research Laboratory in Arizona or at a scheduled World Meteorological
Organization (WHO) inter comparison.  An absolute measurement of the intensity
of the direct beam from the sun is made with an Active Cavity Radiometer.  This
instrument is based on fundamental principles.  The cavity "sees" only the
direct radiation from the sun.   The optically black surface in the cavity is
heated by radiation of all wave lengths.  The cavity temperature Is accurately
measured and the instrument yields the absolute flux of direct radiation  (D) at
the measurement location.  A global pyranometer with a disc located to shield
the direct beams from the disc of the sun measures the diffuse radiation  (d).
With knowledge of the angle of the sun from the zenith (9).  the total global
radiation (G) can be calculated by the following formula.

                    G = D cos 9 •*• d

      Secondary standards,  traceable to such an inter comparison, may be used t
calibrate operational pyranometers.  Pyranometers which have been calibrated in
this way may be used as collocated transfer standards In the field.

-------
                                                             Section No.  4. 1.4
                                                             Revision No.     0
                                                             Date:   17 Sep 89
                                                             Page:    4  of  4
       4.1.4.6 Atmospheric Water Vapor

       It is possible  to  create an  absolute humidity and NBS has a facility for
doing Just that.  Atmospheric  water vapor  instruments, expressing the
conditions as relative humidity or  dew  point temperature can be calibrated by
NBS in a fundamental procedure requiring only the measurement of length
(volume), mass and  temperature.  There  are standard methods for creating
relative humidity environments useful for  some kinds of Instruments.  ASTM
(1985c) describes such a  method.  For air  quality applications, a collocated
comparison provides adequate accuracy.   If the collocated instrument is a
psychrometer and proper methods are used  (ASTM,  1984) for measuring the wet-
and dry-bulb temperatures, traceability to NBS might be claimed for the
thermometers used in the  psychrometer.

       4.1.4.7 Precipitation

       There are some  measurements  where  traceability to NBS is possible but
not required.  Precipitation measurement  is essentially a measure of the volume
of liquid water  (including the liquid water equivalent volume of snow) which is
collected by an area bounded by a cylinder.  Calibration may require volumes of
water or equivalent weights.  The accuracy required and expected from
precipitation gages will  be well served by the accuracy of commercial measuring
equipment.  Ordinary chemical  dispensers  such as graduated cylinders and burets
are accurate enough without calibration traceable to NBS.  The measurement of
the area of  the cylinder  may be made with a commercial ruler or tape, always
keeping in mind  the need  for quality commercial  products for the best accuracy
without extraordinary  effort.

       4.1.4.8  Atmospheric Pressure

       Calibration  labs  create pressures with devices using weights.  It is
possible to  use weights  with calibration traceable  to NBS, but the  accuracy
with which  the atmospheric pressure is  needed for air quality applications does
not require  such  an effort.

-------
                                                             Section No.  4.1.5
                                                             Revision No.     0
                                                             Date:    17 Sep 89
                                                             Page:     1  of  6
4.1.5  ESTIMATING PRECISION AND ACCURACY

       4.1.5.1 Definitions


       There are about as many definitions of precision and accuracy as there
are bodies devoted to carefully defining these terms.  The definition used here
is found in EPA (1976) on page A17 as follows:

"Accuracy - The degree of agreement of a measurement (or an average of
measurements of the same thing). X. with an accepted reference or true value,
T. usually expressed as the difference between the two values, X - T, or the
difference as a percentage of the reference or true value, 100(X-T)/T, and
sometimes expressed as the ratio, X/T."

       For meteorological purposes, this concept of accuracy is acceptable.
The problem comes from knowing the "accepted reference or true value."  Section
4.1.4 Traceabillty Protocol discusses this problem for all the meteorological
variables of Interest.  All data that are
used are averages or means.  The formula for accuracy is


                1   n
           E » 	 V(X, - T )  - X - T                       (1)
    where

      E  is the average error  (accuracy)

      Xt is the ith sample of X

      T  is the non-varying  true value of X
      n  is the number of samples

      i  is a sample, 1,2,3...n

       Accuracy, the average error, or really  the uncertainty  in  the value,  Xj,

has two or three components.  They are bias, conditional bias, and random
error, a statistical expression of a  series of which  is called precision.
Since, in some cases, bias and conditional bias  can be separated, both will  be
discussed.

       Precision is defined  in EPA  (1976) as "A  measure of mutual agreement
aoong individual measurements of the  same property, usually under prescribed
similar conditions.  Precision is most desirably expressed in  terms of the
standard deviation but can be expressed  in terms of variance,  range, or other
statistic.  Various measures of precision exist  depending upon the  'prescribed
sioilar conditions.'"  This  is more difficult  to fit  to meteorological
neasurements made in the atmosphere because "prescribed similar conditions"  are
hard to find.  Parts of the  instrument system  can be  challenged by controlled
environments such as wind tunnels and temperature, humidity or pressure
chambers.  The precision of  the measurement can  be found, providing it is
larger than the variability  of the controlled  environment.  Usually it is not

-------
                                                             Section No. 4.1.5
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    2  of  6


larger and what  Is measured Is the  variability of  the  controlled environment.
The definition to be  used In this handbook is as follows:
    Precision  is  the  standard deviation of  a  series  or measured values,
    X,, about  the mean measured value,  X.   The  foi
    of the precision  or standard deviation, s,  is
X.,  about  the mean  measured value,  X.   The formula for an estimation
                                           2
                                       - X)                     (2)
                                 n - 1
An equivalent  formula which can be used for real-time calculations  in  modern
data loggers or computers (Juran,  1979, p 22-7)  is
                                                  2
                      ± / n  [ (X,2) - (  [ X, )
                                                           (3)
                                  n (n - IT

 Equation  (3)  is preferred because it introduces only two rounding errors rather
 than  the  four introduced in equation (2).  Suppose
 X'  =  X, + C where C is some constant value.  Then

        Xt = X^ - C  and  X = X' - C  and  Xj - X = X^ - Xf.

 Therefore, the standard deviation, s. of a series of values,  as calculated by
 (2) or  (3) is identical to the standard deviation, s, of the same series of
 numbers,  each plus a constant.  If the true value. T, is constant, the
 precision of the accuracy estimate, E. Is identical to the standard deviation
 of the  samples, X,.

        Bias is defined in EPA (1976) as "A systematic (consistent) error in the
 test  results."  It goes on with qualifiers which apply to chemical labs.  For
 meteorology,  the definition for this handbook is simply the average difference
 between the measured value X. and the true value T., (X. - T.).   Using the

 symbols of ASTM (1984),  the  systematic difference (d) or bias is found from the
 following equation.

-------
                                                             Section No. 4.1.5
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    3  of  6
                             n
                               X,  - T,)  = X - T              (4)
             where

               T  is the ith sample of the true value of T


It can be seen In these definitions that when the true value,  T, does not
change,  the accuracy Is the same as the bias.  What is the precision of formula
(D?  For the non-varying true value, the accuracy with which any sample may be
determined is

                     Ei = E ± s                               (5)
and for a variable T

                     Ei = d ± s                               (6)
When T varies, X may vary in a systematic way.  For example,  the case for wind
direction found in 4.2.2.2.2.3 shows a systematic difference,  d, of -3.4 deg.
(orientation error) and a conditional bias (potentiometer linearity error) of
about ± 0.5 deg. on some wave shape with an amplitude of about ± 2 deg.  The
accuracy of the vane might be stated as

                    E. a -3.4 ±2.5 deg.  or, removing the

orientation error,  E. =  0.0+2.5 deg.  or, correcting for linearity

at each angle,      E, =  0.0 ±0.5 deg.  There is little point in
making this final correction in operational programs.  This is just an example
of how a conditional bias can be treated to decrease an error or improve the
accuracy.

       4.1.5.2 Collocated Transfer Standards

       If a collocated transfer standard (CTS) is used to estimate accuracy,
any correction which makes the CTS provide values closer to the true value is
worth making.  The statistical method of operational comparability (ASTM, 1984)
is used when a CTS is employed to estimate the accuracy of an instrument
operating in the field.  The analysis requires differences between the
instrument being challenged and the CTS, properly sited.  The differences are
best if the same data logger can sample the challenged sensor,  X , at the same
                                                                a
tiae as the CTS, X^ so that the difference,  (X .~Xbl)p can be stored, squared

and accumulated.  If this is not possible, the CTS data logger should form
averages over the same time period as the averages formed in the challenged
system.   Then these averages are used for X  , and X. ..  The first calculation
to sake is comparability from the following formula:

-------
                                                             Section No. 4.1.5
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    4  of  6
                   C  = ±  /  	 )   (X_.  - * .)                (7)
             where

                X .  is the ith measurement  of  the subject output

                X.,  is the ith simultaneous sample from the CTS


The systematic  difference,  d, is calculated from (4)  substituting
X . and X. .for  X.  and T,, respectively.  The  estimated standard deviation of
the difference,  s,  is calculated from (8).
                   s = ± /   C2 - d2                           (8)


The minimum  sample size. nQ. required for the calculation of C is given by

equation  (9).   Most data loggers sample sequentially.   The time between members
of a data pair to satisfy the requirement of simultaneous measurements must not
exceed one tenth the response time of the instruments.   The time between pairs
of measurements must be greater than four times the response time of  the
instruments  to assure sample independence.
where
                      no - I—r-l                               (9)
  r is one  increment of resolution reported by X

For example,  a CTS wind vane operating in a speed range of  2  to  7  m/s
with a delay distance of 2 m would have a response time between  1  and 0.3 s.
If a data logger had an analog to digital conversion cycling
time faster than
0.03 s and  if samples were taken no faster than every 4 s,  and if  the
resolution  (measurement and display) of the measuring system  were  1 deg., and
assuming a  5 deg.  standard deviation of the difference,  s.  the minimum  sample
size would  be
    n
     o
225 samples, requiring 15 minutes 9 1  per 4 s.
There cannot  be too many samples.   The minimum is specified  for  a  confidence of
99.7V. or greater that the estimated mean difference,  d.  is within  one  element
of resolution (1 deg. in the example).  At that confidence level,  the  accuracy
of the estimate increases (error decreases) as the square root of  the  sample
size.  If  the sample size were increased by a factor  of  4 to 900,  the  accuracy
of the estimate would by 0.5 deg.   The values of C and d "found from  a  series of

-------
                                                             Section No. 4.1.5
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    5  of  6

differences are only valid for the range of conditions used for the test.  The
shorter the period of time that is sampled the smaller the range of conditions
will be.   A reasonable goal for a minimum CIS time period is 24 hours from the
standpoint of dynamic range variation.

       Assume the CIS will provide the true value in a test in the free
atmosphere within the limits of the calibration of the CIS to some other
standard,  typically 0.1 m/s with respect to some wind tunnel for speed and 1
deg. with respect to TRUE NORTH for average direction.  The accuracy of the
challenged instrument is the comparability, C.  The bias, d, provides the
calibration and orientation error.  The standard deviation, s, provides the
irreducible random error or minimum functional precision with which two
instruments measuring the same quantity report when operated using the ASTM
04430-84,  determines the operational comparability of meteorological
measurements.  Lockhart (1989) found the following values of s for wind speed
and wind direction:

                Wind speed	s a 0.2 m/s

                Wind direction	s a 2 deg.

When s is found to be larger than these values, the assumption must be made
that either site bias or a malfunctioning sensor is at fault.  Under those
conditions the calibration error, d, is also suspect.

       4.1.5.3  Other Considerations

       The average error, E, calculated over a uniform distribution of X. is
the same as the average difference or bias.  The contribution of  the standard
deviation of the difference between X and T (or the precision error) goes to
zero and the average accuracy is the bias.  The measurement of meteorological
variables in the atmosphere is never really that simple.

       Occasionally there is confusion between the word precision and the
resolution of the measurement system.  Resolution is the fineness of the
measurement system, the output of the measurement system or the display of the
output of the measurement system.  A wire-wound potentiometer in a wind
direction sensor may have a resolution of 0.355 degrees  (1,000 windings of a
wire over 355 degrees).  The circuit that converts the resistance of the
potentiometer to voltage may have a resolution of 0.1 degree.  The display or
recorded value may show whole degrees.  The whole degree may be truncated from
the output or rounded.  For example, if the potentiometer wiper is at the 312th
wire (312 x 0.355 = 110.76 degrees), the voltage output  (312/1000 x 1 volt =
0.3120 volts) has no resolution; the resolution of its measurement is limited
only by the resolution of the volt meter.  If the system has a digital output
with a resolution of one degree, the output will be 110 degrees (truncated) or
111 degrees (rounded).  The resolution of the sensor in this example is 0.36
degrees and the resolution of the system is 1 degree.

       A wind direction system with a resolution of 10 degrees might have a
precision of ± 0.5 deg. (it would take a lot of samples to prove  that
precision).  On the other hand, a wind direction system with a resolution of
0.1 deg.  might have a precision of ± 3 deg. (because of hysteresis in the
coupling of the potentiometer to the vane in the sensor). Resolution should be

-------
                                                             Section No. 4.1.5
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    6  of  6


specified to match  the  needs of the data application and to provide sufficient
information for data QC.

       Most of the  discussion in this  handbook includes the sensor and signal
conditioner providing an output.  Good digital data systems degrade the output
so little that they contribute only a  small error  to the total.  The resolution
and accuracy example 4.1.4.1 applies to any digital system.  There is a two
digit uncertainty in the digit which represents the resolution of the
measurement.   If the analog to digital converter resolves wind direction to 10
deg., the accuracy  cannot be better than ± 10 deg.  If the converter resolves
to 1 deg.. rounds to the nearest 10 deg. and displays 10 deg., the accuracy of
the average direction may be 3 deg. while the accuracy of a single observation
is ± 5 deg.

       When analog  recorders are used, their error must be added to the error
of the measurement.  For some recorders this error can get quite large.  Seldom
considered or  specified,  for analog recorders which use rolls of chart paper.
is the error caused by  expansion and contraction of the paper as a function of
temperature and humidity.  This error  added to the resolution uncertainty of
narrow paper rolls  marked by a tapping bar  (when such recorders are used] can
dominate the error  of  the system when  the analog recorder data are used as
measurement data.

       Random  errors identified for each component of. a system can be combined
to estimate the total  system error by  the RSS  (root sum square) method.  Biases
or systematic  errors cannot be combined in this way.  They must be added
arithmetically.  See Fritschen and Gay (1979) for  further discussion of error
analysis.

-------
                                                             Section No. 4.1.6
                                                             Revision No.     0
                                                             Date:   17 Sep 89
                                                             Page:    1  of  2
4.1.6  SYSTEM AUDITS
    A system audit,  as defined in EPA (1976. p A10) is "A systematic on-site
qualitative review of facilities, equipment, training, procedures,
record-keeping, validation, and reporting aspects of a total (quality
assurance)  system, to arrive at a measure of the capability and ability of the
system.   Even though each element of the system audit is qualitative in nature,
the evaluation of each element and the total may be quantified and scored on
some subjective basis."  In short, it is an evaluation of the suitability and
effectiveness of a QA Plan or QA Manual.

    Any audit is most useful when considered as a learning or training
exercise.  Given the newness of the implementation of quality systems to air
quality programs, particularly meteorology, a mechanism for "on the job
training" is useful.  Given the two facts that everyone really wants to do a
good job and almost everyone is a stranger to the concepts of structured
quality systems, an audit is a valuable tool.  There  is really little
difference between air quality and meteorology when it comes to system audit
principles.  The short section on system audits in EPA (1976) is  slanted toward
air chemistry projects.  In a very general way, a system audit should include
the following elements for any technical discipline.

       1.   Declared Agenda - The audit should not be a surprise  or contain
surprises.   The serious audit Is well planned in advance in writing.  The items
to be covered are spelled out.  The agenda  is structured with the help of those
to be audited, recognizing the areas where  they may need special  help.

       2.   Entry and Exit Interview with Top Management - A short introduction
meeting with the authority being audited sets the stage for the cooperation
necessary for success.  Success  is defined  as improving the audited program
through  training and education.  A short exit Interview will announce the
findings already discussed with  the QA people being audited.  The exit
interview and the audit report should contain no surprises.

       3.   Checklist Structure  - The audit  should flow along a prepared
checklist of questions, but if time is  limited as  it  usually Is,  flexibility  is
valuable.  Special  problems, either found or volunteered should be resolved
even at  the expense of failure to finish the checklist.

       4.   Audit Report - The report should be delivered  in a timely manner,
certainly no more than 30 days after the audit, preferably within a week of the
audit.   The report  Is  the  important documentation  verifying the QA program  is
"in control."   It must contain the structure for corrective action with plans
and schedules committed in writing.  An open-loop  pledge or a general plan  is
likely to get a  low priority.  The value of the audit and corrective action
must be  clear  to  the audited organization  if the system audit is  to be
something other  than a paperwork exercise.

       5.   Follow-up  - The QA Plan of  the  organization being audited should
require  some form of documentation of the completion  of the tasks defined in
the corrective  action  plan.  Completion of  this task  closes the loop for  the
system audit.

-------
                                                             Section No.  4.1.6
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    2  of  2


    A short visit to understand  the organization, QA Plans and Procedures, and
to meet the QA person  is  a good  first  step.   It  is difficult to do an effective
system audit without access to all three.

    Since there  is a large spectrum of quality programs which might be audited,
it is important  to keep in mind  the original  reason for the audit and the
objectives and regulations defining the quality  program.  One objective must be
to provide the necessary level of quality at  the minimum  cost as well as to
make the audit process useful and effective.  The goal  is to ensure valid data
with documentation backing up the claim for validity.   This can happen in many
different ways,  some of which may not  conform to the auditor's concept of a
system.  It  is not  the role of the  auditor to redesign  the system, but to
determine if  the-system meets the objectives  and requirements.

-------
                                                     Section No.  4.1.7
                                                     Revision    No.  0
                                                     Date:    11 Aug 89
                                                     Page:     1  of  1
4.1.7     PERFORMANCE AUDITS
     This handbook will concentrate on performance audits.   The audit methods
for each variable will be described in each of the variable-specific sections.
The purpose of a performance audit is to determine as completely as possible
whether  or not the instruments are producing valid data.   It is the
responsibility of the operators to calibrate and operate the instruments so that
they do  produce valid data. As mentioned in Section 4.0.0,  performance audit
methods  may be identical to those used in calibration.  If they are different.
it is expected that the audit method is most comprehensive challenging the
greatest part of the total system.  A complete calibration of an anemometer
requires a wind tunnel.  Most operators do not have access to a wind tunnel and
elect to use the manufacturer's wind tunnel experience as authority for the
anemometer transfer function.  This practice is generally acceptable where
manufacturers can provide test reports confirming their results.  Statistics on
the distribution of error of samples of production run anemometers with respect
to the generic transfer function are necessary if the manufacturer does not
calibrate each unit during manufacture.  If 100% calibration is a part of
manufacturing process, the method of calibration should be available from the
manufacturer.

     A performance audit may include a challenge by a collocated transfer
standard.  Such data serve to check the transfer function at a few points.  The
uncertainty of the challenge is usually greater than wind tunnel challenges
where the conditions can be carefully controlled.

     A performance audit on variables such as relative humidity (or dew point
temperature), solar radiation and atmospheric pressure will usually include a
collocated transfer standard comparison.  This is may be the only way a
challenge can be made  to the whole measurement system.

-------
                                                             Section No. 4.1.8
                                                             Revision No.    0
                                                             Date:   17 Sep 89-
                                                             Page:    1  of  7

4.1.8  DATA VALIDATION PROTOCOL

       4.1.8.1  Strip Charts

       Some years ago, meteorological data were recorded primarily on strip
charts.  Average values were estimated by "reading" the charts.  The most
conmon and the most useful strip chart was one where the output of the
measuring system was continuously recorded by means of a galvanometer movement
or a servomotor.  Data QC meteorologists became expert in determining the
validity of the data by examining the strip chart trace, a process necessary
while checking the digitizing of the strip chart data.

       When the digital data logger and computer first appeared, it was the
recommended practice to have strip chart recorders in parallel to the digital
systems.   The strip chart data could be "read" and used to fill the gaps when
the digital system failed, a common occurance in early designs.  Many
meteorologists found that the strip chart data contained information which was
not present in the digital listing.  The most important information was the
character of the output during the period of time that the digital system was
sampling and averaging.  One example is the presence and frequency of
potentiometer noise in the wind direction output.  This information is an early
indicator'of potentiometer failure.  Whether or not the digital average was
influenced by this noise could be seen by the comparison of the two outputs,
the strip chart being used as truth.

       Another example shows threshold degradation by the character of the
anemometer trace.  Of course the effect of Ice or freezing temperatures on
anemometers and wind vanes could often be seen on the strip chart.  The digital
average value would simply be a number which met the plausibility test but was
erroneous.

       Digital systems have become more reliable, accurate and capable of
on-line processing and at the same time less and less expensive.  The economy
of the digital system pushed the analog recorder to a "back up" role and toward
extinction.  Digital systems show promise of both large memory sufficient to
save one second data samples for time history plots and on-line diagnostic
programs to monitor output patterns for unrealistic variability or lack
thereof.

       Until the technology stabalizes enough to allow a consensus to be
developed and regulatory positions to be taken, requirements for strip chart
data will be a reagon or agency specific requirement.  The one clear fact is
that strip chart data, used as a data quality control tool, will result in a
better data validity protocol.

       4.1.8.2  Methods

       Once data are collected, they should be reviewed to screen out possible
incorrect data points before they are put into accessible storage or passed on
to the user.   While the purpose of a QA program is to avoid the generation of
bad data, it is impossible to do so completely.  Even in the best planned and
best conducted programs,  undetected errors can be generated by faulty
equipment,  noisy data transmission lines, faulty key punching, and a myriad of

-------
                                                             Section No.  4.1.8
                                                             Revision No.    0
                                                             Date:    17 Sep 89
                                                             Page:     2  of  7

other sources.  Filippov  (1968)  offers  a detailed and thorough discussion of
the various possible sources  of  error.

       In both automatic  (AOP) and manual data screening the most obvious
checks should be performed,  first.  These include such things as being sure that
the data exist and are properly  Identified, the forms or files are filled out
properly, that numbers are  in the blocks where they should be, letters are
where they should be. and blanks exist  where nothing should be.  This sort of
data editing is a subject unto itself and will not be pursued here.

       Methods of editing or  screening  meteorological data usually involve
comparison of the measured  value with some expected value or range of values.
Techniques for checking the measured value usually fall into one or more of the
following categories:
       1.   Comparison with upper and/or lower limit on the allowed
            range of the  data;
       2.   Comparison with known statistical distribution of
            the data;
       3.   Comparison with spatial and/or temporal data fields; and
       4.   Comparison based  upon known physical relationships.

       A choice must also be  made of what to do with the datum that does not
pass a validation procedure.   Basically there are two choices, eliminate the
questionable data from the  file, or flag it for further examination.
Automatically discarding  data may be a  viable, cost-effective option if the
screening procedure is carefully designed and each datum is not of high value.
Records must be kept of discarded data  so the reason for the fault can be found
and corrected.  Flagged data  are examined and a decision made on their
acceptability.  If unacceptable, it may be possible to correct them or
substitute a more reasonable  value  (Reynolds, 1979).  Corrected or substituted
values should be so indicated in the data file, with an explanation of the
substitution available to the user.  Alternatively, data of questionable value
may be kept in the data file  under a flagged status, with a notation of why
they are questionable, so that the user can make a decision as to their
usefulness.  This procedure^ is of questionable value to most users because the
collecting agency is frequently  In the  best position to make a decision on the
data.

       The range test is  the  most common and simplest test.  Data are checked
to see if they fall within  specified limits.  The limits are set ahead of time
based usually upon historical data or physically impossible values.  Some
examples of reasonable range  tests are  rainfall rate greater than 10 in./hour
or wind direction not between 1  and 360 .  In setting the limits,  one must
take into consideration whether  or not  the system will select only outrageous.
extreme  (i.e., impossible)  values usually caused by data handling errors (such
as wind speeds greater than 100  m/s or  less than zero) or Just unusually high
(i.e.. possible ) values, which  should  be examined further.  This may require a
further decision on Just  how  extreme a  value should be flagged.  This decision
should be based on the real Impact of using extreme values should they be in
error.  Considerations of the cost of incorrect data, the possibility of
correction or substitution, or replacement by obtaining new data should be
made.  Unfortunately, the decision may  also frequently be made on the available
resources of those who examine the flagged data.

-------
                                                             Section No. 4.1.8
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    3  of  7

                4.1.8.2.1  Comparison with known distributions

                Comparison with known statistical distributions may involve
comparison of means, standard deviations, means of extremes, or higher order
statistics.  For example, Lee and Stokes  (1978) report that their data base
usually had kurtosis of approximately 3 with zero skewness.  Any of their
instruments, then,  that showed a marked departure from these values were
considered to be In need of further verification.  (Additional research is
needed to determine whether these or similar criteria could be used in other
areas.)

                Lockhart (1979) suggests  compressing data  into a densely packed
graph where long-term  (week, month, or seasonal) patterns  can be easily seen.
Major departures from  these subjectively  seen patterns can be noted and the
data checked.  Although this method of data verification is usually used to
check a particular  data set against a longer term climatology, it can also be
used to check Individual values.  For example, one might compare a temperature
reading with the monthly average maximum  or minimum plus or minus
(respectively) two  or  three standard deviations.  This technique obviously
depends on a reliable  history or representative measurements being available
from the site and is Ineffective for noting significant long-term changes in
the instrument.

                4.1.8.2.2 Comparison with other data fields

                Screening data by comparison with fields of similar or related
data is commonly done  when large amounts  of data are taken and when assumptions
of spatial continuity  of the meteorological variable are physically reasonable.
The most easily visualized example of this Is a field of atmospheric pressure
measurements.  Any  value can be compared  with those in a large area around it.
either visually, or by numerical interpolation.  Major deviations from the
dominant pattern (a low pressure reading  In the middle of  a high pressure area)
are not to be expected.  Of course, allowance must be made for meso- and micro-
scale phenomena such as a shortwave or pressure Jump area  ahead of a convective
storm.

                Not all meteorological fields can be expected to have the
needed continuity.  Rainfall is a notorious example of discontinuity or
microscale variations.  Wind speed and direction can exhibit continuity on some
spatial scales, but care must be taken to account for the  many effects, such as
topography, that can confuse the issue (See Section 4.0.4.3.2.4).

                Interrelated fields can also be used to screen data.  Rainfall.
for example, is unusual without clouds and high humidity while wind direction
and speed, especially  above the surface layer, are related to pressure
gradients.

                Fields of data in time, rather than space, are also used to
check datum points.  These checks are usually made on rates of change of the
data.   Checks are made both on rates of change that are too highland not high
enough.   For example,  atmospheric stability is not expected to change by
several classes within an hour.  A wind direction reading, however, that does
not change at all for  several hours may indicate that the  vane is stuck

-------
                                                              Section No.  4.1.8
                                                              Revision No.    0
                                                              Date:   17 Sep 89
                                                              Page:    4   of 7


(assuming the wind speed is not zero) or that there is some other problem with
the system.

                 4.1.8.2.3 Comparison based on physical relationships

                 Screening checks can also be made to assure that physically
Improbable situations are not reported in the data.  This kind of check  is not
commonly used because of the wide variety of conditions that can occur in the
atmosphere under extreme conditions.  These unusual events would frequently be
noted first  by  some of the statistical or range checks noted above.

                Table 4.1.8.1  Examples of Data Editing Criteria
      Wind Speed:
      Wind  Direction:
      Delta Temperature:
>25 m/s (NRC)
>50 kts (NCC)
>20 kts and doubles at 3-hour observation (NCC)
First 5 hourly values <±0.2 mph of next 4 (TVA)
Any recorded calm wind speed (NCC)
Same sector for more than 18 hours (NRC)
First 5 hourly values <± 2° of the next 4 (TVA)
                                and 5 p.m.  (TVA)
                                and 5 a.m.  (TVA)
10
   a.m.
     m.
      Stability:
       Temperature:
AT/Az >1 C/lOOm between
AT/Az <-l°C/100m between 6 p
AT/Az >15°C/100m (TVA)
AT/Az <-3.4°C/100m (autoconvectiveMTVAMNRC)
AT/Az changes sign twice in 3 hours (TVA)
A.B.F, or G stability during precip.  (NRC)
F or G stability during the day (NRC)
A.B. or C stability during the night  (NRC)
Change in stability of more than 3 classes
     between 2 consecutive hours (NRC)
Same stability class for >12 hours (NRC)
9°F > mean dally maximum for the month (TVA)
9°F < mean daily minimum for the month (TVA)
> 10°F change in 1 hour at a site (TVA)
First 5 hours within ±0.5*F of next 4 (TVA)
>125°F (NCC)
<-60°F (NCC)
> 10°F change 1
                                           hour or 20 F in 3 hours (NCC)
       Dew Point:
       Pressure:
Dew point >  temperature (TVA)(NRC)
Dew point change >7°F in 1 hour (TVA)
First 5 hours within ±0.5°F of next 4 (TVA)
>90°F  (NCC)
<-60*F  (NCC)
Temp. - dew  point >5°F during precip. (NRC)
Temp. = dew  point >12 consecutive hours (NRC)

>1060 mb  (sea level) (NCC)
<940 mb  (sea level)  (NCC)
Change of 6  mb  or 0.2 inch Hg in 3 hours (NCC)

-------
                                                             Section No.  4.1.8
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    5  of  7


                Some screening points of this type that are used include
assuring that the dew point is not greater than the temperature, and that the
lapse rate is not greater than the autoconvective lapse rate.  Checks on
stability class versus time (not allowing "strongly unstable" at night or
"stable" during the day) may also be considered in this category.

                Table 4.1.8.1 gives examples of some of the data editing
criteria used by three Federal agencies:  the National Climatic Center (NCC,
now NCDC), Kllnt (1979); the Nuclear Regulatory Commission (NRC), Falrobent
(1979);  and the Tennessee Valley Authority (TVA), Reynolds (1978).  Examination
of the table shows some interesting differences that can be ascribed to  the
differing missions of the agencies.  Because of their global concerns, the NCC
must allow a far wider range of limits on fields such as temperature and
humidity than does an agency with only local Interest, such as TVA.  On  the
other hand, the NCC has the data available to do spatial checks over a wider
area than would be possible for many local study situations.  Differences can
also be noted depending on the type of data collection (spot readings once per
hour or three hours versus continuous recordings) and major interests (synoptic
weather patterns versus stability).  Flllppov (1968) gives an exhaustive review
of checks used by weather services of many other countries.  The criteria
listed in the table are used to Identify data to be edited or challenged for
further review.

       4.1.8.3 The AREAL System

       On the following page is a data validation system recommended for AREAL
to replace the present system for screening meteorological data.   It could be
used to screen data gathered by AREAL, contractors, or state and  local
agencies.  The system takes into account the variable nature of AREAL's  field
activities.  It does not depend on, or have the advantages of,  long-term
multistation network design, nor is it labor intensive.  The basic goal  is that
of rapid identification of field problems, with low value assigned to
individual data points, thus allowing the discard of questionable values.
Flexibility is available, however, if an individual project's meteorological
data are judged to warrant a more critical approach.

       The flow of the system is shown in Figure 4.1.8.1.  All data will go
first through a hard copy auditing procedure designed to find data entry and
keypunch errors.  In the hard copy audit, a percentage of data points will be
randomly selected for audit.  A second, independent file of  these values, as
well as the hour just before and after the hour, will be created  from the
original hard copy.  This file will be compared with the master file and
discrepancies noted.  If there are only a few random discrepancies, these
points will be eliminated from the system.  If there are several, or there
seems to be a systematic pattern of errors, the project office  (the office
responsible for gathering and reducing the data) will be notified  so that they
can correct and re-enter the data and correct the data entry system.  The data
are next passed through a screening program, which is designed  to note and flag
questionable values.  Flagged data will go to the laboratory meteorological
office for review.  There they will either be accepted, discarded, or returned
to the project officer  if there is a large amount of questionable data.  That
officer may accept, discard, or correct the data.  The screening  values  are
given in Table 4.1.8.1.  They offer a combination of range,  rate  of change, and
physical impossibility checks that are chosen to be reasonably  restrictive.   It

-------
                                                                Section No.  4.1.8
                                                                Revision No.     0
                                                                Date:    17 Sep 89
                                                                Page:     6  of  7
is anticipated that some good data will  be flagged, but that most data handling
and gross  instrument failure problems  will be caught.
      Stan
                                   J
r
Edit/ Merge
Program
I




Correct Data
Acquisition/
Processing
Procedures
                          Umalidated
                             File
                   Compansion of
                   Selected Data
                 with Data Obtained
                 Independently from
                   Original Source
  Project Office
Determine if Suspect
  Data are to be:
    Approved.
    Corrected.
    or Deleted.
                  Are Bad Data
                  Indicative of
                   Recurring
                   Problem?
                                    Flagged
                                    Data
                      Meteorological
                  Screening Program
                      Determine if Data
                        Approved.
                        Corrected.
                        or Deleted.
                                                            Unresolved
                                                            Data
                   Comparison of
                   Subset of Data
                  withN.W.S. Data
            Questionable
            Data
                                             Data Corrections.
                                             Approvals, or
                                             Deletions.
      Figure 4.1.8.1
Schematic flow of decisions in the
 AREAL data validation scheme.
        Data that pass the  screening program will go through a comparison
program.   This  program will randomly select certain values for manual
comparison with information collected by  the National Weather Service.   In the
selection process, one day and one hour will be chosen on which  data from all
stations in the network will be audited.   One day in every 20 will  be randomly
chosen, and on  that day, one hour between 5 and 9 a.m. (EST), will  be randomly
chosen.  Data from that hour and day for  all stations will be printed out by
the audit program for the  manual checks.   The program will also  make compressed
time scale plot (20 days of hourly values on one line) for each  parameter for
the use of the  validators.

        The data generated  by the audit programs will first be compared with
National Weather Service data to see if  they fit reasonably well with synoptic

-------
                                                             Section No.  4.1.8
                                                             Revision No.     0
                                                             Date:    17 Sep 89
                                                             Page:     7  of  7


conditions prevalent in that area.   The meteorologist will choose the stations
to be used in the verification,  and train the data clerks in the subjective
comparison procedure.   All questionable data will be given to the meteorologist
for review as above.  The variables to be checked In this way will Include wind
speed and direction, temperature,  dew point, pressure, and occurrence of
precipitation.

       Naturally, if the audit checks show a problem with one or more
instruments, an attempt will be made to Identify the time range of that problem
so that all questionable data can be found.  Logs of bad data will also be kept
and used to identify troublesome instruments and other problems.

       This system is suggested principally for AREAL, but may prove a useful
starting place for state and local air pollution agencies wishing to develop a
meteorological data validation procedure.  The suggested system Is very
complete and will be evaluated over a period of time.  Changes to the system
may have to be made, depending upon the needs and resources of the users.

-------
                                                             Section No. 4. 1.9
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    1  of  2

4.1.9  QA REPORTS AND CORRECTIVE ACTION

       4.1.9.1  Operations Log and Maintenance Reports

       In all of section 4.1 it has been stressed that the real purpose for a
QA or Quality program Is to document data validity and the steps taken to make
that determination.   Any activity which has the potential of affecting the
validity of the data must be reported.  A report usually includes a note in the
station log indicating the time of the operator visit and visual status of the
sensors.   The log is signed by the operator.  If the operator found a problem
which he was authorized to fix. the log would contain the entry to that effect.
If the operator is not authorized to make the repair or does not have the
necessary parts, a maintenance report can initiate the work and the purchase
of the parts to fix the instrument.  When parts are changed, serial numbers or
property numbers should be noted in the log.  If a part does not have a number
(sone cup assemblies and propellers do not have numbers), a number of some sort
should be assigned and marked on the part with permanent ink.  The QA Plan
should provide some communication route and method by which the person
responsible for the project and the person responsible for data quality control
(if they are different people) are notified of work done on the system.

       4.1.9.2  Calibration Reports

       Calibration reports are the most crucial documents of a data collection
project.   They are the foundation blocks which uphold the the validity claim.
Quality Control and the routine inspection of the data spans the time between
calibrations.  The calibration reports will show whether or not the system is
"in control."  If the system is always "in control" or operating within the
required tolerance limits stated in the QA Plan and generated by the
application, and the data QC does not have any unsolved mysteries, the data are
valid.  If the calibration shows problems, the report will also show the
corrective action taken or initiated.  The "as-found" and "as-left" readings
are a vital part of the calibration report.  If any data "correction"
(quotation marks used because this is a very delicate subject) is applied, the
Justification must rest on calibration reports on either side of the data
period and the data in between.  This report requires distribution to the
project leader and the QC inspector, or at least a sign-off routing.   If
corrective action is initiated but not completed, a report of completion is
required and has the sane routing.

       4.1.9.3  Audit Reports

          Audit reports should confirm the calibration reports.  If they do
not. the assumption is that the audit report is correct.  Whenever a
measurement discrepancy exists, the cause of the difference must be found and
resolved.   If the audit measurements are wrong,  the auditor will be smarter
next time and all parties will have more confidence in the calibration reports.
It is the responsibility of the auditor to include a report of the discrepancy
between calibration data and audit data along with the explanation and solution
of the discrepancy in the audit report. It is the responsibility of the
operator to be sure that it is in the report.  The documentation must be
suitable for use in a court of law.

-------
                                                             Section No.  4.1.9
                                                             Revision No.     0
                                                             Date:   17 Sep 89
                                                             Page:    2  of  2
       4.1.9.4  Reports  to Management
       Reports to management  are of value  to maintain the close communication
necessary between the source  of authority,  top management, and the exerciser
delegated authority, the QA organization.   Whatever  the structure of the
organization, directed  effort must be  paid for and planned for through
budgeting.  Top management must know how the quality program is performing its
intended, money saving  role.   There are about as many types of meteorological
aonitoring programs as  there  are applications.  One  fairly standard one is the
2-level, 60 m tower used at most nuclear power plants.  Crutcher (1984)
provides an insight into costs of a minimum system and an acknowledgement of
the annual costs involved in  operations and quality  assurance.

"Costs are controlled by the  design and reliability  of the system, as well as
the marketplace.  Costs given here are approximate 1977 prices for presently
available equipment sufficient to meet the minimum requirements of the Nuclear
Regulatory Commission's Regulatory Guide 1.23  (formerly the USAEC Safety Guide
23).  For the first two years these costs  approximate one-third of a million
dollars.  These minimum costs do not  include either  office or storage space.

    One  tower,  installation and equipment                    $100,000
    Annual maintenance  cost                                   25.000
    Annual cost of surveillance and quality assurance
        (including personnel and supplies,
       magnetic tape, paper etc.)                              50,000
    Annual cost of data listings, etc. based on  15-min.
        Integrating intervals  and automatic logging in
       digital  form on  magnetic tape,  13 parameters
        (channels) to a  page,  daily summaries                   60,000

The cost of a mobile tower and equipment installation is essentially 50% of the
cost of  a permanent-type tower installation.  Other  costs remain essentially
the same."

    Of particular interest in this reasonably accurate estimate  (ten years ago)
is  the  ratio of annual  costs  to one-time costs,  1.35:1.  Of course, the
requirement of  RG 1.23  is for valid data with documentation and quality
assurance.  Smaller systems cost  less, but the often neglected provision for
annual  operating and QA costs are  still necessary  If valid data are required.

        4.1.9.5  Discrepancy Reports

        Some systems report discrepancies as a  section of another report and
soae use a discrepancy  report as  a stand-alone vehicle to initiate corrections
and report completed corrections.   If it is a  stand-alone report, some system
of  control  is necessary to keep track of open  reports and monitor progress
toward  completion  (called follow-up or needling).

-------
                                                             Section No.  4.1.10
                                                             Revision No.      0
                                                             Date:     17 Sep 89
                                                             Page:      1  of  2
4.1.10.      REFERENCES
ANSL/ASQC,  1987a:   Q90. Quality Management and Quality Assurance Standards -
    Guidelines for Selection and Use. Araer. Soc.  For Quality Control.
    Milwaukee. WI  53203

AKSI/ASQC,  1987b:   Q91, Quality Systems - Model for Quality Assurance in
    Design/Development, Production, Installation, and Servicing.  Amer.  Soc. For
    Quality Control,  Milwaukee. WI 53203


ANSI/ASQC,  1987c:   Q92. Quality Systems - Model for Quality Assurance in
    Production and Installation. Amer. Soc. For Quality Control, Milwaukee. WI
    53203.

ANSI/ASQC.  1987d:   Q93. Quality Systems - Model for Quality Assurance in Final
    Inspection and Test.  Amer. Soc. For Quality Control,  Milwaukee, WI 53203.

ANSI/ASQC.  1987e:   Q94. Quality Management and Quality System Elements
    Guidelines.  Amer.  Soc. For Quality  Control. Milwaukee, WI 53203.

ASTM.  1984a:  Standard Test Method for MEASURING HUMIDITY WITH A
    PSYCHROMETER (THE MEASUREMENT OF WET- AND DRY-BULB TEMPERATURES.
    E.  337-84.  Amer. Soc. for Testing and Materials, Philadelphia, PA 19103.

ASTM.  1984b:  Standard Practice for DETERMINING THE OPERATIONAL COMPARABILITY
   OF  METEOROLOGICAL MEASUREMENTS, D4430-84.  Amer. Soc.  for Testing and
    Materials, Philadelphia. PA 19103.

ASTM.  1985c:  Standard Practice for MAINTAINING CONSTANT RELATIVE HUMIDITY BY
    MEANS OF AQUEOUS SOLUTIONS. E 104-85.  Amer.  Soc. for Testing and Materials.
    Philadelphia.  PA 19103.

Crosby, P.  B. . 1979:   Quality is Free.  McGraw-Hill Book Co.,
    ISBN 0-07-014512-1.

Crutcher, H. L., 1984:   Monitoring. Sampling, and Managing Meteorological
    Data, Chapter 4.   Atmospheric Science and Power Production, Darryl
   Randerson. Ed.   DOE/TIC-27601, pp. 136-146.

EPA. 1976:   Quality Assurance Handbook for Air Pollution Measurement Systems,
    Vol.  I, Principles, EPA-600/9-76-005.  Office of Research and Development,
    3es.  Triangle Park, NC 27711.

EPA. 1987b:  On-Site Meteorological Program Guidance for Regulatory Modeling
    Applications.  EPA-45O/4-87-013, Off. of Air Quality Planning and
    Standards. Res. Triangle Park. NC 27711.

Fairsbent,  J. E. ,  1979:   (personal communication) Nuclear Regulatory Commission.

Feigenbaum, A. V., 1961:  Total Quality Control Engineering and Management.
    McGraw-Hill, LCCCN 61-11051

-------
                                                             Section No. 4.1.10
                                                             Revision No.     0
                                                             Date:    17 Sep 89
                                                             Page:     2  of  2

Filippov, V. V. ,  1968:   Quality control procedures for meteorological data.
    World Weather Watch Planning Report,  No.26,  World Meteorological
    Organization.

Fritschen, L.  J.  and L.  W.Gay,  1979:   Environmental instrumentation.
    Springer-Verlag,  NY,  ISBN 0-387-90411-5.

Grant, E. L  .  and R.  S.  Leavenworth,  1974:   Statistical Quality Control, Fifth
    Edition.   McGraw-Hill,  ISBNO-07-024114-7

Juran. J. M. ,  1979:   Quality Control  Handbook,  Third Edition.   McGraw-Hill,
    ISBN 0-07-033175-8

Klint, W. E. ,  1979:   Screening checks used by the National Climatic Center for
    meteorological  data.   NCC Asheville,  NC (unpublished)

Lee, J. T. and J. Stokes,  1978:  Tall tower and aircraft instrumentation quality
    control  procedures - development  and application.  Proceedings 4th
    Symposium  on Meteorological Observations and Instrumentation.  Aner.
    Meteor.  Soc., pp 3-6.3-6.

Lockhart, T. J.,  1978b:   Data graphics for assessment of measurement quality.
    Proceedings of  the National Conference on Quality of Environmental
    Measurements, Denver.  Nov.  27-29, pp 123-132.

Lockhart. T. J. ,  1985c:   Quality assurance of meteorological measurements.   In:
    "Quality Assurance for Environmental Measurements," ASTM STP 867.
    J. K. Taylor and T.  W.  Stanley (eds). Amer.  Soc. for Testing and Materials,
    Philadelphia. PA. 253-259.

Lockhart, T. J. ,  1989a:   Comments on "A Quality Control Program for Surface
    Mesonetwork Data".  J.  Atmos. and Oceanic Techno1., 6, 525-526.

Lockhart, T. J. ,  1989b:   Accuracy of the collocated transfer standard method
    for  wind instrument auditing.  J. Atmos. Oceanic Techno1.,  6, 715-723.

Reynolds, G. W.  and D. E.  Plttman. 1978:   The TVA meteorological data acceptance
    analysis program.  Proceedings 4th Symposium on Meteorological Observations
    and  Instrumentation. Amer. Meteor. Soc., pp 3-6.3-6.

Reynolds, G. W. ,  1979:  Final acceptance review for TVA meteorological  data.
    Presented  at Quality Assurance in Air Pollution Measurement Conference.
    Air  Pollution Control Association, New Orleans.

Wade. C. G. ,  1987:   A quality control program for surface mesometeorological
    data.  J.  Atmos. and Oceanic Techno 1., 4, 435-453.

-------
                                                             Section No. 4.2.0
                                                             Revision No.    0
                                                             Date:  Sep  17 89
                                                             Page:    1  of  4


                                  Section 4.2
               QA  FOR WIND SPEED,  WIND DIRECTION AND TURBULENCE
                                    OUTLINE
Section                                                    Pages  Rev.   Date
4.2.0  OUTLINE AND SUMMARY                                   4     0     9/89

4.2.1  TYPES OF INSTRUMENTS                                  4     0     9/89

       4.2.1.1  COMMON MEGHANICAL SENSORS
            4.2.1.1.1  Wind Speed Sensors
            4.2.1.1.2  Wind Direction Sensors
            4.2.1.1.3  Fixed Component Sensors

       4.2.1.2 SECONDARY EFFECT SENSORS
            4.2.1.2.1  In-situ Sensors
            4.2.1.2.2  Remote Sensing Devices

    4.2.2  SPECIFICATIONS                                   32     0     09/89

       4.2.2.1  WIND SPEED
            4.2.2.1.1  Threshold
                4.2.2.1.1.1  Definition
                4.2.2.1.1.2  Threshold Measurement
                4.2.2.1.1.3  Starting Torque Measurement
            4.2.2.1.2 Accuracy
                4.2.2.1.2.1  Definition
                4.2.2.1.2.2  Measurement of Accuracy
                4.2.2.1.2.3  Application of Accuracy
                4.2.2.1.2.4  Precision
            4.2.2.1.3  Distance Constant
                4.2.2.1.3.1  Definition
                4.2.2.1.3.2  Measurement of Distance Constant
            4.2.2.1.4  Off-Axis Response

       4.2.2.2  WIND DIRECTION
            4.2.2.2.1  Threshold
                4.2.2.2.1.1  Definition
                4.2.2.2.1.2  Threshold Measurement
                4.2.2.2.1.3  Torque Measurement
            4.2.2.2.2  Accuracy
                4.2.2.2.2.1  Definition
                4.2.2.2.2.2  Measurement of Sensor Accuracy
                4.2.2.2.2.3  Measurement of Orientation Accuracy
                4.2.2.2.2.4  Expression of Accuracy
                 4.2.2.2.2.5  Precision
            4.2.2.2.3  Delay Distance (Distance Constant)
                4.2.2.2.3.1  Definition
                 4.2.2.2.3.2  Measurement of Delay Distance
            4.2.2.2.4  Overshoot  or Damping Ratio
                4.2.2.2.4.1  Definition
                 4.2.2.2.4.2  Measurement of Overshoot
                4.2.2.2.5  Dynamic Vane Bias

-------
                                                             Section No.  4.2.0
                                                             Revision No.    C
                                                             Date:  Sep   17 89
                                                             Page:    2   of  4


Section                                                    Pages  Rev.    Date
       4.2.2.3  TURBULENCE
            4.2.2.3.1  Definition
            4.2.2.3.2  Direct Measurement of Sigma Theta
            4.2.2.3.3  Statistical  Summaries

4.2.3  ACCEPTANCE TESTING                                    5     0      9/89

       4.2.3.1  WIND SPEED
            4.2.3.1.1  Threshold
            4.2.3.1.2  Accuracy
            4.2.3.1.3  Distance Constant
       4.2.3.2  WIND DIRECTION
            4.2.3.2.1  Threshold
            4.2.3.2.2  Accuracy
            4.2.3.2.3  Delay Distance and Overshoot
       4.2.3.3  MEASUREMENT SYSTEM

4.2.4  INSTALLATION                                          8     0      9/89

       4.2.4.1  GENERAL  CONSIDERATIONS
       4.2.4.2  WIND SPEED
       4.2.4.3  WIND DIRECTION
            4.2.4.3.1  Exposure
            4.2.4.3.2  Orientation
                4.2.4.3.2.1  True Solar Noon Method
                4.2.4.3.2.2  Solar  Azimuth Method

4.2.5  CALIBRATION                                           4     0      9/89

       4.2.5.1  WIND SPEED
            4.2.5.1.1  System Accuracy
            4.2.5.1.2  Component  Accuracy
       4.2.5.2  WIND DIRECTION
            4.2.5.2.1  System Accuracy
            4.2.5.2.2  Component  Accuracy

4.2.6  OPERATIONS.  MAINTENANCE  AND QUALITY CONTROL           2     0      9/89

       4.2.6.1  OPERATIONS
       4.2.6.2  MAINTENANCE
            4.2.6.2.1  Routine  and Preventive  Maintenance
            4.2.6.2.2  Corrective Maintenance
       4.2.6.3  QUALITY  CONTROL

4.2.7  PERFORMANCE AUDIT METHODS                            20     0     9/89

       4.2.7.1  GENERAL  CONSIDERATIONS
            4.2.7.1.1  Who
            4.2.7.1.2  What
             4.2.7.1.3  Where
             4.2.7.1.4  When
             4.2.7.1.5  How

-------
                                                             Section  No.  4.2.0
                                                             Revision No.     0
                                                             Date:  Sep  17 89
                                                             Page:     3  of  4


Section                                                    Pages  Rev.    Date
       4.2.7.2  WIND SPEED
            4.2.7.2.1  Sensor Control Method
            4.2.7.2.2 CIS Method
            4.2.7.2.3 "W" Anemometers
       4.2.7.3  WIND DIRECTION
            4.2.7.3.1 Sensor Control
            4.2.7.3.2 CIS Method
            4.2.7.3.3 Vertical Wind Direction, 
       4.2.7.4  TURBULENCE OF 
-------
                                                             Section No. 4.2.1
                                                             Revision No.    '.
                                                             Date:  Sep  17 85-
                                                             Page:    4  of  4


                QUALITY ASSURANCE FOR WIND SPEED, WIND DIRECTION
                                 AND TURBULENCE
                                     SUMMARY

     This section discusses  all  aspects of the  task of monitoring the wind at =
particular site with an emphasis on quality assurance.  A background chapter
describes the nature of wind and the kinds of instruments commonly used  to
monitor  its  speed and direction.  This section  describes in detail the
important aspects of the operation of conventional anemometers  and wind  vanes.
Some discussion of  secondary effect sensors is  provided but the handbook is nc.
intended to  cover these instruments.  The background  information and the
detailed information found in the following chapters  are necessary for two
kinds of tasks.  One is to execute the responslblity  for the collection  of
valid data.  The other is to audit or Judge how well  the first  task was
performed within the goals or regulations which caused the measurements  to be
made in  the  first place.

    Specifications  is the longest and in some ways the most detailed section.
The premise  is  that an understanding in depth of the  way the common sensors
work is  necessary before purchasing,  installing and operating the instruments.
Specifications  set  the performance parameters for the instrument or system.
Careful  definitions are given along with test methods which will equip the user
to verify or to judge the work of others who verify confonnance to
specification.
    Once the specifications  are clearly understood, the process of purchasing
and acceptance  testing can be considered.  The  contention is that quality
assurance is a  vital aspect  of defining that which is to be purchased and
verifying the performance of the delivered system.  When the valid system is Lr
hand, the installation can be planned and Implemented.  The important process
of orientation  of the wind vane to TRUE NORTH  is described  in detail.

    Calibration is  a foundation on which claims of data validity are built.
This important  function may  be practiced  in a number  of phases  of the
monitoring program.  This chapter stresses documentation of the calibration
findings and methodology.  The use of the most  inclusive methods practical in
field conditions  is advocated.  Once the calibrated system has  been installed,
the routine  performance of operational checks,  preventive and corrective
maintenance  and quality control operations begin.  The documentation of  these
operations provide  the framework, resting on  the foundation of  calibration,  tc
support  the  claim of data validity.

    Performance audits add confidence to  the documentation  that the system is
in control.  Performance audit methods must be  the most comprehensive Methods
possible for field  challenges.  This chapter describes some recommended  audit
methods  and  audit forms to support the methods.  The  performance audits  and
calibrations provide the data for the estimation of accuracy and precision
described  in the  following chapter.

    There  is not  room to include all the details or background  information
which might  be  needed or desired.  A list of  the references used for this
section  is  found  at the end of  the section.   If the reader  needs additional
information  or  is curious about peripheral  subjects,  the references will
provide  the  answers or a start  in search of  the answer.

-------
                                                             Section No. 4.2.1
                                                             Revision No.    ;
                                                             Date:    17 Sep 85
                                                             Page:     1  of  4
4.2.1  TYPES OF INSTRUMENTS
    There are many ways to detect wind as it passes a point on Earth.  Only
those ways which reference a fixed point (or volume) will be considered in this
handbook.  This class of measurements is expressed in Eulerian coordinates
where properties of the air are assigned to points in space at each given tine
(Huschke, 1970).  The other class of measurement is expressed in Lagrangian
coordinates.   It is good to keep in mind that Eulerian measurements are
frequently used in Lagrangian models.  Or,  in other words,  monitoring data
measuring wind on a tower are used to estimate where parcels of air move and
how the concentrations of constituents of the parcel change in the process.

    It is necessary to understand Just how the measurement is made to
adequately do the following:
       o write procurement specifications,
       o adopt and apply acceptance testing methods,
       • site the sensors in the representative flow of interest,
       • perform calibration and maintenance services,
       e establish an effective quality control (QC) operation, and to
       « evaluate audits used to estimate precision and accuracy of the data.

    This section will describe how various kinds of instruments work.  The
thoroughness of the description in this handbook will be proportional to the
frequency of use of the instrument in air quality applications.

    Another background point deals with the nature of wind.  There can be no
question about the wind requiring a vector to describe fully a single
measurement.   The vector has direction in spherical coordinates (azimuth with
respect to TRUE NORTH and elevation with respect to a horizontal plane) and
length (speed) along that direction.  It is common for many air quality
applications to deal only with the presumably horizontal flow as measured by
vanes and anemometers.  In this case, the horizontal components of the vector
are expressed as an azimuth angle from which the wind is blowing and the speed
at which the air is passing the point of measurement.   While each sample of
wind requires both speed and direction, it is common to measure them
separately.   A series of samples may be summarized in different ways depending
upon the application.  The arithmetic mean of the samples is recommended for
many applications (EPA. 1987b).  The standard deviation of the samples Is used
to describe the level of turbulence in the air.

       4.2.1.1  Common Mechanical Sensors

                4.2.1.1.1  Wind speed sensors

                Common anemometers are either cup assemblies turning on a
vertical axis or propellers turning on a vane-oriented horizontal axis.  The
cup anemometer is an empirical sensor in that the relationship between the rate
of rotation and the wind speed is determined by testing rather than defined by
theory.  It is a linear relationship, for all practical purposes, above Its
threshold non-linearity and through the range of important application.  It is
an aerodynamic shape which converts the wind pressure force to torque (hence
rotation) because of asymmetrical lift and drag.  Its dynamic performance
characteristics (starting threshold and distance constant)  are density

-------
                                                             Section No. 4.2.1
                                                             Revision No.    I
                                                             Date:   17 Sep 6?
                                                             Page:    2  of  4

dependent but  its  transfer function (rate of rotation vs. wind speed) is
independent of density.   The cup is not very efficient and creates turbulence
as the air flows through and around it.   The linear  speed of the center of a
cup Is only a half to  a  third the linear speed of  the air turning it (Mason and
Moses, 1984).  The cup anemometer is omnidirectional to horizontal flow but
exhibits a complicated reaction to vertical  components.  It may indicate speed
slightly greater than  the total speed when the flow  is non-horizontal
(MacCready, 1966).

                The propeller anemometer is  a more efficient shape.  The
helicoid propeller is  so efficient that its  transfer function can be specified
from theory (Gill,  1973).   It creates little turbulence as the air flows mostly
through it. turning like a nut on Invisible  threads.  The propeller measures
wind speed when it Is  oriented into the wind by a  vane.  Its errors from
imperfect alignment with some mean vector are small, being nearly proportional
to the cosine of the angle of misalignment.

                In either of these types of  anemometers, the rate of rotation
is sensed by some  transducer.   Tachometer generators, a.c. frequency
generators, light  choppers and shaft revolution counters have all been used.
It is important to know  how the transducer works if  the performance of the
anemometer is to be challenged for a QA purpose.

                4.2.1.1.2  Wind direction sensors

                The wind vane is perhaps the simplest of instruments.  A fin is
tied to a vertical shaft such that when force is applied to the area by the
wind, it will turn the shaft seeking a minimum force position.  The
relationship of the shape,  size and distance from  the axis of rotation of the
fin to the bearing assembly and transducer torque  requirements determines the
starting threshold.  These attributes of the fin area along with its
counterweight determine  the dynamic performance characteristics of overshoot
(damping ratio) and delay distance (distance constant) of the direction vane.
While its equilibrium  position is insensitive to density, the dynamic response
characteristics and threshold are density dependent.

                Vane design is of little importance  if the average wind
direction is all that  is required.   If turbulence  parameters are of interest,
as they usually are or should be,  the design of the  vane becomes important.
The vane transducer is usually a potentiometer,  but  synchros, shaft encoders,
capacitors and Hall effect devices have been used.   It is fairly common to find
theorange of the sensor  to be "540 degrees"  rather than the physically true
360 .  The reason  is related to the problem  of a continuous range (a circle)
with a discontinuous output (0 to n volts).   It is important to know how the
transducer works if the  performance of the wind vane is to be challenged for a
QA purpose.

             o  A  special direction vane is  the bivane which has the vertical
range of ± 45  to  60 in  addition to the full azimuth circle.  The additional
range brings with  it the need to neutralize  gravity  by having a perfectly
balanced vane assembly.   Blvanes can be conditionally out of balance, such as
happens when dew forms and then evaporates from the  tail fins.  The effect of
this Imbalance on  threshold and performance  is complicated.  Horizontal vanes

-------
                                                             Section No. 4 2.1
                                                             Revision No.    I
                                                             Date:    17 Sep £=•
                                                             Page:     3  of  4

can be designed to be stable in the horizontal even when slightly out of
balance.  The effect of this design is to add the vane horizontal restoring
force to the wind force, again a complication.


                4.2.1.1.3 Fixed component sensors

                Propeller anemometers exhibit something like a cosine response
to a wind along some line other than the axis of the propeller.  The degree
with which this response represents a cosine is a function of the design of the
propeller.  If the cosine response is perfect, the fixed propeller accurately
reports the component of the wind parallel to the axis of the propeller.  If
three propellers were located on fixed X, Y and Z axes, the three outputs would
define the components of a three dimensional wind vector.  From a QA
perspective, the accuracy of the wind speed and direction data are related to
the determination of the component errors and the algorithms used to correct
for them.  Often ill-defined are the errors from the interference of one
propeller on another and the errors when the beyond-threshold-nonlinearity
speed has not been reached.  It is the speed of the component parallel  to the
anemometer axis that the propeller responds to, not the total speed.  A 5 m/s
wind with a 5° up angle and 5° off the Y axis will provide a 0.44 m/s wind for
the W and X propellers.  A 50% error in the X propeller because of threshold
nonlinearity would cause an Insignificant 0.014 m/s error in the wind speed and
a 2.5  error in wind direction.  A SO'/, error in the W propeller for the same
reason would cause a 50% error in the W component (0.22 m/s reported rather
than the true 0.44 m/s).

       4.2.1.2  Secondary Effect Sensors

                4.2.1.2.1  In-situ sensors

                Several meteorological instrument books contain information on
a variety of wind instruments.  See Mason and Moses (1985) and Middleton (1953)
for greater depth and variety.

                The three component sonic anemometer is considered in some
circles as the standard for wind measurement.  For those applications where the
contribution of small eddys is important, it is an excellent choice.  As with
many of the secondary effect sensors, it is a research tool requiring
considerable attention from the operator.  It is not a good choice for  routine
monitoring.  It has its own set of error sources when it is used for measuring
long-term (tens of minutes) averages and standard deviations.  Single component
sonic anemometers deployed as W sensors may become the standard for this
difficult measurement.  The secondary effect for this instrument is the
transport of sound waves in the air.

                The hot wire or hot film anemometer is also a research  tool
which measures the wind speed component perpendicular to a heated cylinder.
The secondary effect for this instrument is the removal of heat by the  air
measured by the current it takes to replace the heat.  There are some new
designs of this type of instrument which are intended to be monitoring
instruments.

-------
                                                             Section No. 4.2.1
                                                             Revision No.    0
                                                             Date:    17 Sep 89
                                                             Page:     4  of  4
                4.2.1.2.2 Remote sensing devices
                There are two Doppler shift  instruments which measure wind
remotely by analyzing the return from transmitted  energy pulses.  The most
important for  the  boundary layer applications is the acoustic Doppler (SODAR)
sounder.  It depends on the back scatter from small temperature differences
that tag the air for motion measurement.   The other Doppler uses
electromagnetic energy to measure winds through the troposphere.  There are
also some Doppler  applications using lasers  as the energy source. ,  These
systems are complicated.   A QA effort related to systems of this type will
require special study of the system and ingenuity to find other ways of
measuring what they are measuring.  In the case of the SODAR which is being
used in monitoring applications for air quality programs, a reasonableness test
with another instrument capable of measuring winds aloft is more likely to
produce useful information than is a challenge designed to report the accuracy
of the SODAR.

-------
                                                             Section No. 4.2.2
                                                             Revision No.    j
                                                             Date:   17 Sep 89
                                                             Page:    1  of 32
4.2.2  SPECIFICATIONS
    The purpose of defining specifications is to give unambiguous meaning to
the terms used by all those who are concerned that the instruments and systems
selected and operated will meet the needs of the application or project.  This
starts with procurement specifications and ends with supporting claims of data
quality. These specifications provide the basis for receiving inspection and
testing.  The wind is the most important variable to be measured and its
specifications are the most complicated.  Specifications discussed here will
also include some aspects of the measurement system.

    Project and application requirements vary.  To make this handbook as
specific as possible, the examples used will be consistent with those presented
in the On-Slte Meteorological Program Guidance for Regulatory Modeling
Applications (EPA, 1987b).  The specifications will be discussed in order of
their importance and then summarized at the end of the sub-section.

       4.2.2.1  Wind Speed

                4.2.2.1.1  Threshold

                     4.2.2.1.1.1 Threshold definition

                     One of the keys to a good wind sensor is a low threshold.
The threshold Is also the one performance characteristic which will certainly
change with time because of bearing degradation.  There is no standard
definition of threshold so different manufactur-ers may apply different tests
to establish their threshold specifica-tion.  Absence of a standard or
definition of the specification makes it difficult to specify a meaningful
value.  The following definition comes from Standard Test Method for
DETERMINING THE PERFORMANCE OF A CUP ANEMOMETER OR PROPELLER ANEMOMETER (Draft
6) (ASTM, 1985):
         "Starting threshold (U ,m/s)—the lowest  wind  speed  at  which  a
          rotating anemometer starts and continues to turn and produce
          a measurable signal when mounted in its normal position."
A starting threshold specification, 0.5 m/s for example, should include a
footnote describing the meaning of the specification. In the example above. It
might say:  0.5 m/s (1)
         (1) "as determined by wind tunnel tests conducted on production
          samples in accordance with ASTM D22.ll test methods."

                     All rotating anemometers are non-linear as they go from not
turning to turning at a rate predicted by their linear transfer function.
Note that the definition does not require linear output at threshold,  only
continuing turning and measurable signal.  If the manufacturer provides an
accuracy specification which is independent of speed, the presumption is that
the accuracy specification is met at threshold.  Consider a hypothetical cup
with a transfer function. I.e., the relationship between rate of rotation and
wind speed, as follows:
                                U • 0.2 + 1.5 R
                        where U Is wind speed (m/s) and
                          R is rate of rotation (rps)

The transfer function would have been found by using a least squares fit
(linear regression) to wind tunnel data.  The ASTM method uses the wind speeds
well above the starting threshold to avoid bias from the non-linear threshold.
In Figure 4.2.2.1 the lowest 2 m/s of the hypothetical performance curve is

-------
                                                             Section  No.  4.2.2
                                                             Revision No.    0
                                                             Date:    17 Sep 89
                                                             Page:     2  of 32
shown along with the contribution of the offset to the system output.   The
variable part of the transfer function (U = 1.5 R) coining from the  cup rotation
is shown theoretically as the straight line from 0.2 m/s to an output  of 1.8
m/s when the wind speed is 2 m/s.  The triangles show the actual  output from
the cup rotation.  They start to turn at 0.3 m/s (threshold) and  reach the
theoretical line at about 0.8 m/s.  The parallel line through the origin simply
adds the constant offset to the cup rotation output.  The measurement  error is
the difference between the diamonds in the figure and the ideal straight line.
It starts at +0.2 m/s, goes to -0.1 m/s at 0.3 m/s, and then gets smaller as
the nonlinearity of the threshold decreases.

                     The offset is defined either by the linear regression or
by the arbitrary choice of the manufacturer.  If it is the former,  the starting
threshold will always be larger than the offset.  If it is the latter, the
starting threshold may be either side of the offset.  The manufacturers of the
common small three cup anemometer often set an offset voltage in  their signal
conditioner as shown In Figure 4.2.2.1.  For this hypothetical cup,  the offset
voltage is critical to its meeting the accuracy specification discussed in
4.2.2.1.2.  Sensitive propeller anemometers have a much smaller offset because
they develop more force (torque) per m/s. Some offsets are so small that there
is no advantage or need to use an offset voltage.  See Baynton (1976)  and
Lockhart (1977) for further discussion of the errors of rotational  anemometers.
particularly at the threshold.

                Cup  Anemometer Performance
                                 Threshold Analysis
       2.0

       1.8 H
   3
   O
Hypothetical Transfer Function
U (m/s) B 0.2 + 1.5 R (rps)
              From Cup Rotation and Offset Voltage

                      Ideal

               O     Actual
                                              From Cup Rotation

                                                      Theoretical

                                                V    Actual
                 0.2     0.4    0.6    0.8   1.0    1.2    1.4

                                  Wind Speed  (m/s)
                                                1.6
Figure 4.2.2.1   A hypothetical cup anemometer threshold  analysis.

-------
                                                             Section  No.  4.2.2
                                                             Revision No.     0
                                                             Date:    17 Sep  89
                                                             Page:     3  of  32
                     4.2.2.1.1.2   Threshold Measurement

                     There  is  only one way to measure starting threshold.   It
requires a wind tunnel capable of  accurate operation below  1 m/s. One  standard
methodology is defined in ASTM (1985) and described in Lockhart  (1987).
However, it is possible  to  estimate the starting threshold  by matching the
torque which is required to keep a cup or propeller from turning at a  known
wind speed (in a wind tunnel)  with the starting torque of the anemometer
bearing assembly.  Lockhart (1978) provided the torque relationship as a
function of wind speed for  four anemometer shapes.

                     Table  4.2.2.1 contains values calculated with these
data by using the relationship
                            T = K u
          where T  is torque (g cm /sec )
                u  is the square of the wind speed (m/s)
                K  is a  constant for the aerodynamic shape  (g)
The values in the table  were calculated from this formula using
the K values from Lockhart  (1978).

                Table 4.2.2.1  - Torque Developed vs.  Wind Speed
Speed
(m/s)
0.1
0.2
0.3
0.4
0.5
1.0
cup
#1
(g-cm)
0.014
0.056
0.126
0.224
0.350
1.4
cup
#2
(g-cm)
0.027
0.108
0.243
0.432
0.675
2.7
prop
#3
(g-cm)
0.049
0.196
0.441
0.784
1.225
4.9
cup
#4
(g-cm)
0.148
0.592
1.332
2.368
3.700
14.8
#1 Teledyne Geotech 170-42 (20.3 g) K= 1.4
#2 MRI Model 1022 (48.3 g) K= 2.7
#3 R.M. Young Model 21180 (9.7 g) K= 4.9
«4 MRI Model 1074 (186.8 g) 10*14.8
The torques listed are those acting on the sensor when the sensor Is restrained
in a wind field at the speed listed.  If the sensor bearing assembly has a
starting torque less than the torque provided at that speed, and the restraint
is removed, it will start turning.  The torque watch used for the low speed end
of the wind tunnel tests was a Waters Model 366-3 with a range from 0.003 oz-ln
(0.216 g-cm) to 0.03 oz-in (2.16 g-cm).   To convert oz-ln to g-cm,  multiply by
72.
                     The method of using a starting torque measurement to find
the sensor starting threshold will become standard only with the publishing of
K constants by the manufacturers.  One manufacturer (R.M. Young Co.) provides
the K value for anemometers.  These values are shown in Table 4.2.2.2.

-------
                                                             Section No. 4.2.2
                                                             Revision No.    C
                                                             Date:    17 Sep 8?
                                                             Page:     4  of 32
                      Table 4.2.2.2 - Anemometer K Values
Type
Polypropylene Cupwheel
Polypropylene Propeller
Polystyrene Propeller
Polystyrene Propeller
Model
No. 12170C-100cm
No. 08234- 18x30cm
No. 21282-19x30cm
No. 21281-23x50cm
K
1.4
2.5
3.6
5.0
                     4.2.2.1.1.3 Starting Torque Measurement

                     The starting torque of an anemometer bearing assembly will
increase in time because of wear and dirt.   The starting torque, with the cup
assembly or propeller removed, can be measured. Starting torque measurement is
simple in concept but sometimes difficult in application.  An experienced
meteorological instrument technician can tell if a bearing assembly is in need
of service by simply feeling the shaft or rotating or spinning the shaft and
listening to its sound.  The trouble with this practice is that it is not
quantitative.   It works for field servicing instruments but does not provide
documentation suitable for a quality control program.  Another qualitative
practice is to roll the sensor slowly over a smooth horizontal surface watching
                                            the shaft not turn as the sensor
                                            turns around it (see Figure
                                            4.2.2.2).   Set screws and other
                                            asymmetries apply a torque which
                                            keeps the shaft from turning while
                                            the sensor moves around it.  If th«
                                            applied torque could be measured,
                                            this method would be quantitative.

                                                  The measurement of the
                                            starting torque of the bearing
                                            assembly provides only an
Figure 4.2.2.2 Climatronics F460 torque     approximation of the starting
               test for speed sensor        threshold of the anemometer,
particularly cup anemometers.  The absence of the cup weight may lower
the starting threshold of the cup bearing assembly but there is no evidence
that this is an important consideration.  At this point in time there is no
better way to estimate and document in the field and in units of wind speed
this important specification, the starting threshold of the anemometer.

  The direct measurement of starting torque requires some device which can
apply a known torque.  The most common, perhaps, is the Waters Torque Watch.  A
model 366-3 is shown in Figure 4.2.2.3 applied to a Climatronics cup anemometer
sensor.  The measurement requires some degree of care and skill.  The torque
watch has a square shaft which fits into a square hole in the connecting
fixture.  The torque watch is turned while holding its shaft in line with the
anemometer shaft, without end loads.  The indicator is watched and when the
shaft turns the maximum reading is recorded.  This process needs to include at
least one full turn of the anemometer to be sure the maximum friction in the
bearing assembly is encountered.  The torque watch measures either clockwise or
counterclockwise.  Use only the rotation sense required by the cup assembly or

-------
                                                             Section No.  4.2.2
                                                             Revision No.     0
                                                             Date:    17 Sep  89
                                                             Page:     5  of  32
propeller. The  range of  the  torque watch may not be as sensitive  as one would
like.  If the model 366-3  is used on anemometer #1 in Table 4.2.2.1,  the
threshold (0.003  oz-in or  0.216 g-cm) will not measure equivalent speeds below
0.4 m/s.  If the  torque  watch turns the shaft
without reaching  the lowest  scale point, all
that can be said  is that the starting threshold
of the anemometer is less  than 0.4 m/s.
     Another torque watch  is  the GM-CM Torqmeter
781 with a 0.1  to 2.1 g-cm range (0.001-0.029
oz-in), shown in Figure  4.2.2.4 mounted to a
Teledyne Geotech  1565C  wind direction sensor.
A third torque  measuring device is the simple
Torque Disc, model 18310 made by R.M. Young Co.,
shown in Figure 4.2.2.5. This is a fundamental
device which does not need expensive
calibration.  Weights (screws) are attached at
distances from  the center  of  rotation.  The
force of gravity provides  g-cm torques at the
center of rotation of the  intentionally out of
balance disc.   The shaft being tested must be
horizontal and  symmetrical in mass.  A cup
anemometer shaft which does not turn while the
sensor is slowly rolled  along a flat surface
will not work with the Torque Disc.  The g-cm
torque applied  equals the  weights and distances
when the weights are in  the same horizontal
plane as the shaft.  Calibration results from
weighing the weights and measuring the distances.
         Figure 4.2.2.3 Waters
                       Torque Watch

       An appropriate  interface
fixture would allow  the Torque Disc to be used to calibrate a torque  watch.
Figure 4.2.2.4 Gm-Cm
Figure 4.2.2.5 Young Torque Disc

-------
                                                             Section No. 4.2.2
                                                             Revision No.    0
                                                             Date:   17 Sep S?
                                                             Page:    6  of 32
 There are several ways to measure torque but the available
instrumentation to make the measurement is limited.   Figure 4.2.2.6 is a
collection of spring-type torque watches, spring scales and circular discs
capable making torque measurements within narrow ranges and specific
orientations.  It is necessary to become familiar with these devices and how
they are correctly used.

                                                         A - Waters Torque Watch
                                                             Model 366-3
                                                             (0.2-2.0 g-cm)
                                                         B - Waters Torque Watch
                                                             Model 651X-3
                                                             (18-360 g-cm)
                                                         C - GM-CM Torqmeter
                                                             Model 781
                                                             (0.1-2.1 g-cm)
                                                         D - Young Torque Disc
                                                             Model 18310
                                                             (0.1-15 g-cm)
                                                         E - Haldex AB
                                                             Gran Gauge
                                                             (1-10 g)
                                                         F - Young Gran Gauge
                                                             Model 18330
                                                             (0-10 g)
Figure 4.2.2.6  Various Torque Measuring Devices

                4.2.2.1.2 Accuracy

                     4.2.2.1.2.1 Definition

                     The classic definition of accuracy is the comparison of a
measured value to a true value expressed as a bias term plus or minus a randoa
uncertainty (precision).  The bias term may be conditional with respect to the
best fit straight line; it may vary with wind speed or angle of attack.

                     When accuracy is specified, the kind of true value to be
used to test the accuracy claim must also be specified.  Usually the buyer
expects the "true value" to be the wind speed where the anemometer is sited.
The manufacturer expects the "true value" to be the near laminar flow of a
calibration wind tunnel.  Some auditors expect the "true value" to be the
output predicted by the transfer function when the anemometer is rotated at a
known rate.  Let us label the kind of accuracy as follows:
       A(l) - accuracy with respect to the horizontal component
              of wind speed at the sited location,
                Ada) - instrument response
                A(lb) - siting representativeness
       A(2) - accuracy determined in a wind tunnel, and
       A(3) - accuracy of conversion of rate of rotation to output.
                     A(3) is the easiest to measure and represents most of  the
claims for data accuracy from audit reports.   It requires, usually, a measure

-------
                                                             Section No. 4.2.2
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    7  of 32


of the offset voltage  (or equivalent m/s) and the combination of offset and
output from the rate of rotation to output converter.

                A(2) requires a calibrated wind tunnel and it is a more
difficult and more expensive accuracy to determine.  It does provide a check on
the manufacturer's generic transfer function plus the variation in production
individuals represented by the individual being tested.  The wind tunnel
calibration test provides a specific transfer function which can be used to
change the signal conditioning to get the smallest error of an A(2) type for
that specific system.  Serial numbers should be recorded with the test results.

                     Some manufacturers do not identify their cup assemblies or
propellers with numbers.  Assume an A(2) accuracy is found In a wind tunnel
test and compared to the generic transfer function used for A(3) operations.
Assume the test shows  that the A(3) value Is uniformly 10% low.  Assume  the
operators took a year  of data with A(3) accuracy tests showing insignificant
errors.  The error to  the A(2) level Is a 10% bias and the data can be
corrected for the year.  This action requires either evidence or good reason to
believe that there was no physical change to the cup assembly over the year and
that the individual tested was the one used during the year.

                     There are still unknowns of the A(l) type to consider.
These are usually conditional biases and often impossible to define.  They may
be recognized and their impact estimated.  There are two types of these  errors.
One Is the consequence of the anemometer design In the flow field It is  to
measure., Ada).  The  other Is a result of assumptions of representativeness,
                      The discussion of A(la) errors requires an understanding
of conventions for  the  use of u, v and w.  In traditional diffusion
applications, the statistics for wind representing a period of time refer  to u
as the speed of  the horizontal component along the direction of the mean wind.
v as the speed of the horizontal wind component perpendicular to the mean
direction, and w as the speed of the vertical wind component.  Another
convention applies  to fixed component anemometers such as the UVW propeller
array.  Here, the U is  the east-west component of the wind in a Cartesian
coordinate system (a west wind Is positive); V is the north-south component of
the wind (a south wind  is positive); and W is 'the vertical component of the
wind (upward' moving wind is positive) (Stull, 1988).

                      MacCready (1966) characterized errors in anemometers  when
operating in a turbulent flow.  A cup anemometer has a u-error because of  a
different response  "constant" to an Increasing speed than to a decreasing
speed, so-called overspeedlng.  With modern sensors, this is usually a small -
error of a percent  or two, depending on sensor design and height above ground.
A cup anemometer has no v-error since it is insensitive to changes In
direction, but it does  have a w-error caused by non-horizontal flow.  This
error can easily be 10% and larger with some designs (Lockhart, 1987).  A  vane
oriented propeller  will have small v-errors and w-errors from misalignment.
These will be small because the propellers respond nearly as the cosine of the
misalignment angle,  2%  for a 10 degree misalignment. The u-error is too small
to measure for light weight helicoid propellers.  These are all Ada) errors
and they vary as the wind varies.

                      A(lb) errors deal with the assumption is that the
anemometer is measuring what the true wind would be at the point of measurement

-------
                                                             Section No. 4.2.2
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    8  of 32


if the anemometer were not there.   This is a question of representativeness and
not instrumentation but it can have a large impact  on the question of data
accuracy.  The  influence of any supporting structure can bias the flow which
the anemometer  faithfully measures.  If the assumption goes further to equate
the measurement to  Its physical height above ground, and if the anemometer is
mounted on a  2  m pole on top of a large 8 m building, the bias with respect to
a 10 m flow over a  flat field will be the fault of  the building.  These errors
are of the A(lb) type.  If the pole is on the edge  of the building, the
distortion of the building will provide non-atmospheric errors of the Ada)
type to be combined with the A(lb) type.  These types of errors are very
difficult to  define and virtually impossible to correct.  Data from an
anemometer mounted  in a questionable site, after the A(2) and A(3) errors have
been calibrated out, could be compared with data from a vane oriented propeller
anemometer mounted  In a space where the subject anemometer  is assumed to
represent.  The difference in these collocated measurements may be used  to
estimate  the  magnitude of AC1) errors.  The A(2), A(3) and A(la)  errors  are the
ones to concentrate on minimizing.  See 4.2.4.1 for siting guidance.

                      4.2.2.1.2.2  Measurement of accuracy

                      The accuracy of an anemometer is found by comparing its
output  to the known speed in a wind tunnel.  A calibrated wind tunnel has
uncertainties associated with Its operation.  These inclvide  Instrumentation
errors  in measuring the wind speed In the tunnel when  It  is empty (0.1  mph or
0.05 m/s  in the NBS wind tunnel above 2 mph) and the  inhomogeneity of wind
speed  In  the  test  section away from the boundary layer  (a function of the
tunnel  design).  The  turbulence level In  the wind tunnel  test  section should  be
homogeneous across  the test section with  most of the  energy in eddy  sizes which
are small compared  to  the size of  the anemonreter.  When this  is  true, and it
usually is,  turbulence does not Influence the calibration.   Fluctuations in  the
tunnel  speed can be thought of as  long wave  length longitudinal  turbulence.
This  "turbulence"  can Influence the calibration without careful  measurement
synchronization and time averaging.

                       When an anemometer  is placed in the test  section for
calibration,  consideration must be given to  blockage errors,  which are
dependent on the ratio of  the  size of  the instrument to the size of  the test
section.   Also  interference errors, which are dependent on the placement of  the
anemometer with respect  to  the wind  tunnel  instrumentation need to be
considered.  Small  calibration wind  tunnels  may  themselves be calibrated with
an anemometer  which has  been  run  in  the NBS  wind tunnel.   It is also a  common
practice to  run two anemometers side  by side, one of which has an NBS
 calibration  curve.   It is  prudent  to  reverse the positions from time to time to
 verify test  section homogeneity.  It  is not  reasonable to expect such
 calibrations,  even though they are "traceable  to NBS" by some definition, to
 have an accuracy better  than  0.2  mph (0.1 m/s).

                       While a  calibrated wind tunnel is the recognized standard
 method for calibrating an anemometer,  a fundamental (but not very practical)
 calibration  is possible by moving an anemometer  over a measured  length in a
 measured period of time through  still air (Lockhart. 1985b and Stearns. 1985).
 Most manufacturers have samples  of their products  calibrated by NBS to
 establish for  their design a generic relationship  between wind speed and rate
 of rotation, measured by counting pulses, frequencies or output voltages.  This

-------
                                                             Section No.  4.2.2
                                                             Revision No.    j
                                                             Date:    17 Sep 89
                                                             Page:     9  of 32
relationship is then used as the transfer function to define what the signal
conditioning electronics or other output devices require to express the
measured rate of rotation in units of wind speed.  Some manufacturers test and
adjust each cup wheel or propeller to fit the generic relationship within some
error band.
                     Baynton (1976) discusses the calibration of anemometers
and shows the results of tests of 12 different kinds of anemometers in the
National Center for Atmospheric Research (NCAR) wind tunnel.  He compares his
calibration to the manufacturer's calibration or generic relationship.  Except
for the Aerovane, which probably was too large for the NCAR tunnel, the
difference was within ±3%.  He also discusses the difference between an
anemometer transfer function which goes through the origin, of the form
                              Y - bX,
and the transfer function with an offset or threshold, of the form
                              Y = a + bX.
Table 4.2.2.3 summarizes data from his Table 1 and Table 5.

           Table 4.2.2.3  Wind  Tunnel  Test  Results From Baynton
a b
Type of Anemometer (m/s) (m/rev. )
Gill 4-blade hellcold propeller 0.073» 0.309
Gill 3-blade hellcold propeller 0.011* 0.487
Aerovane helicoid propeller 0.233 1.356
Taylor Biram' s propeller 0.145 0.255
Case 1 la Sensitive 0.467 1.404
Thornthwaite 0.331 1.476
INSTAAR t 0.316 1.597
Cllmet 011-1 0.265 1.382
TechEcology t 0.275 1.391
Gill 3-cup 0.250 1.057
Electric Speed 0.610 2.728
Bendix Totallizer Model 349 0.588 2.605
MRI Model 1074 tt 0.087* 2.314
* a is not significantly greater than zero
t Institute for Arctic and Alpine Research, Boulder,
* Analysis of NBS data provided by L. Petralll
tt Data from Lockhart (1977)
E

1.03
1.03
0.93
1.00
0.98
0.97

1.01


1.03
0.97


Colo.


E is the ratio of the test result to the mfg.'s calibration.
Baynton  lists the value of "a" for  the  12 anemometers he tested, which ranged
from nearly zero (0.01 m/s) to 0.6  m/s.  One purpose of his paper was to
caution  users of "wind run" instruments of  the  errors associated with ignoring
"a."  Wind run  is generally used  to describe those anemometers which count
shaft rev-olutions over a long period of time or are geared to a counter  or
recorder in such a way that the output  is in units of speed.  These instruments
cannot pro-vide an offset.  Lockhart  (1977) shows that some cup designs,
specifically the Meteorology Research,  Inc. Model 1074 used on the Mechanical
Weather  Station, can  be nearly as accurately described without an offset  (a=0)
as with  one (a-0.03).  Table 4.2.2.4  lists  the  NBS data and the linear
regressions to  sup-port this fact.  The residual errors from each model are
plotted  in Figure 4.2.2.7.

-------
                                                                     Section No.  4.2.2
                                                                     Revision No.     0
                                                                     Date:    17  Sep 89
                                                                     Page:    10   of 32
             Table 4.2.2.4   NBS Test  Data for MR I Model  1074

Test
No.

1
2
3
4
5
6
7
8
9
10
11
12
13
14
IS
14
17
18
Test date
	 NBS-

Output Tunnel
Freq.

-------
                                                             Section No. 4.2.2
                                                             Revision No.    Q
                                                             Date:    17 Sep S9
                                                             Page:    II  of 32


                     A similar analysis for a propeller anemometer is shown :-
Table 4.2.2.5.   These data come from a test in the Atmospheric Environment
Service (AES) of Canada wind tunnel on a propeller anemometer being used in a
"round robin" experiment to estimate the accuracy of wind tunnel calibrations.
Each test was run for 100 seconds.  The tunnel speed is an average of one
second samples taken every ten seconds by AES.  The sensor count is a total for
100 seconds from the light chopper delivering 10 pulses per revolution.  Each
test was replicated and tests 3 and 4 were also replicated in tests 19 and 20.
Two linear regressions were run. The first, and best fit, allowed the intercept
of the X axis,  or zero offset in ASTM language, to be calculated.  The second
forced the straight line through the origin.  This latter method yields a
constant slope or pitch (meters per revolution) which when multiplied by the
rate of rotation (revolutions per second.) results in wind speed  (meters per
second).  The residualerror from these two regressions are plotted in Figure
4.2.2.8.

                     It is characteristic for helicold propellers to show a
better correlation with wind tunnel speeds than does a cup anemometer. This is
because propellers generate torque uniformly without sensitivity of position.
Three-cup assemblies, on the other hand, produce three peaks and three valleys
in torque for each revolution (Lockhart, 1985).  Either type of anemometer can
be calibrated to an accuracy sufficient for most applications.

                     4.2.2.1.2.3  Application of accuracy specifications

                     An accuracy specification should include enough
information  to define the type of accuracy intended and the method by which
accuracy claims may be tested.  Here are a few examples of accuracy
requirements.
                     In the Ambient Monitoring Guidelines for Prevention of
Significant  Deterioration (PSD)  (EPA, 1987a),  it states that for horizontal
wind systems "Wind speed systems  should be accurate above the starting
threshold to within 0.25 m/s at speeds equal  to or less than 5 m/s.  At  higher
speeds, the  error should not exceed 5 percent of the observed speed  ( maximum
error not to exceed 2.5 m/s)."   In the On-Site guide (EPA, 1987b) in 8.1.1  it
states  "Accuracy (error) (1) (2)             s(0.2 u/s +5% of observed)
          (1) as determined  by wind tunnel  tests conducted on production
             samples in accordance with ASTM  D-22.11 test methods. 21  (sic)
          (2) aerodynamic shape  (cup or propeller) with permanent serial
              number to be  accompanied by  test report, traceable to NBS.
              showing rate  of rotation vs. wind speed at 10 speeds."
By Implication, the  latter  specification refers to accuracy type A(2).  although
the expectation is that A(l) will be  Included by careful siting.  This
expectation  must be addressed with experienced subjective judgment.

                     Assume a system  is  to be used  in accordance with EPA
(1987b),  and an "off the shelf"  anemometer is purchased.  The manufacturer
states  that  the sensor delivers  30 pulses  per revolution (ppr) with a  transfer
function  from revolutions per second, R  (rps),  to wind speed, U  (m/s).  of

          U  (m/s) = 0.224 *  -!L°- -  0.244 *
                  =  0.224 (m/s)  + 1.410 (m/r)  R (rps).

-------
                                                       Section No. 4.2.2
                                                       Revision No.    0
                                                       Date:   17 Sep 89
                                                       Page:   12  of 32
             Table 4.2.2.5  NBS Test Data for Young 27106 at AES
	 AES 	
Test
No.

1
2
3
4
5
6
7
6
9
10
11
12
13
14
15
16
17
18
19
20
Output
Count
(1)
3660
3675
10839
10911
13108
13207
25418
25433
33006
33003
40163
40139
47636
47577
55102
55105
62613
62785
10930
10931
Tunnel
Speed
(•/S)
1.14
1.14
3.27
3.27
5.42
5.42
7.59
7.61
9.82
9.83
11.96
11.95
14.13
14.14
16.35
16.30
18.53
18.56
3.29
3.28
(I) 10 pulses oer
Ataosoneric
Y
Tunnel
Speed
(•/s)
1.14
1.14
3.27
3.27
5.42
5.42
7.59
7.61
9.82
9.83
11.96
11.95
14.13
14.14
16.35
16.30
18.53
18.56
3.29
3.28
I
Output
/1 000
(rps)
3.66
3.68
10.89
10.91
18.11
18.21
25.42
25.43
33.01
33.00
40.16
40.14
47.64
47.58
55.10
55.11
62.61
62.79
10.93
10.93
	 Linear fsaression 	

r
(i/S)
1.15
1.15
3.28
3.29
5.41
5.44
7.57
7.58
9.81
9.81
11.92
11.92
14.13
14.11
16.33
16.34
18.55
18.60
3.29
3.29

X'-Y
(l/S)
0.01
0.01
0.01
0.02
-0.01
0.02
-0.02
-0.03
-0.01
-0.02
-0.04
-0.03.
0.00
-0.03
-0.02
0.04
0.02
0.04
0.00
0.01

X1
U/S)
1.09
1.09
3.23
3.24
5.37
5.40
7.54
7.55
9.80
9.80
11.92
11.91
14.14
14.12
16.35
16.36
18.58
18.63
3.24
3.24

X'-Y
(l/S)
-0.05
-0.05
-0.04
-0.03
-0.05
-0.02
-0.05
-C.Oi
-0.02
-0.03
-0.04
-0.04
0.01
-0.02
0.00
0.06
0.05
0.07
-0.05
-0.04

Regression Output:
Constant
Std Err of Y Est
R Squared
Nc. of Observations
Degrees of Freedom

X Coefficient (s)
Std Err of Coef.



Regression Output:
Constant
Std Err of Y Est
R Squared
No. of Observations
Degrees of Freedoe

I Coefficient (s)
Std Err of Coef.


No
0.
0.
0.



X'
. 1-20
067574
023965
9999S4
20.00
18.00

0.295224
0.000279





X1
No. 1-20
0.000000
0.043528
0



0
0

.999944
20.00
19.00

.296803
.000268

revolution counted for 100 seconds.
Environoent Service of Canada
April 12.
1983 «ith
J. Earle
Chaoaan, Young 27106 £ 10/rev.
               Linear Regression of AES-RR Data
U.3 -
0.4 -
0.3 -
0.2 -
*~x
V)
— 0.1 -

2
ii] -0.1 -
-0.2 -

-0.3 -

-0.4 -
O.5 -







9 *














"» ^ 	 *














	 	 ,-—"*-













l 	 J
==f 	


KEY

V X"
	 X'

            15             10            15
                              Wind Speed (m/s)
Figure 4.2.2.8  Residual Errors from Helicoid Propeller
19

-------
                                                             Section No. 4.2.2
                                                             Revision No.    0
                                                             Date:    17 Sep 89
                                                             Page:    13  of 32
                     Assume that an NBS test was -conducted after a year of
operation and the results provided a least squares analysis of
         U (m/s)  = 0.301 (m/s) + 1.387 (m/r) R (rps).
What action is suggested by this finding?  Throughout  the year the operator had
the electronics trimmed to output 0.224 m/s when the cups were not turning and
14.325 m/s when the cups were turning at 10 rps.   The A(3) error in converting
R to U is 0.00. The A(2) error can be expressed as follows:
            Generic transfer function:  U  =  0.224 + 1.410 R
                  Wind tunnel (truth):  U' =  0.301 + 1.387 R
     The error (E. m/s) is U - U' or    E  = -0.077 + 0.023 R
     In terms of the measured speed, U, E  = -0.081 + 0.0163 U
Table 4.2.2.6 compares this error with the specification at different rates of
rotation.  The 1.6 percent overestlmation of speed by the generic transfer
function is not large enough to bother with data correction.  The data meet the
accuracy guidelines with two thirds of the allowable error unused.  At the next
calibration the system should be adjusted to the wind tunnel derived transfer
function.
                        Table  4.2.2.6   Wind  Speed Errors
R
(rps)
0.000
1.225
3.388
6.993
14.203
U
(m/s)
0.224
1.951
5.001
10.084
20.250
U*
(m/s)
0.301
2.000
5.000
10.000
20.001
E
(m/s)
-0.077
-0.049
0.001
0.084
0.249
Allowed
(m/s)
±0.20
±0.30
±0.37
±0.45
±0.81
Used
C/.)
38
16
0
19
31
                     4.2.2.1.2.4  Precision

                     The definition of accuracy describes a bias term and a
variable term akin to precision.  Traditionally, precision describes the
uncertainty with which a measuring process or instrument realizes the measured
value when that being measured  is the same thing and is repeatedly measured.
The key to finding the variability of a measuring process is to use a
non-varying subject.  In meteorology, and particularly in anemometry, it is not
possible to have, with certainty, a non-varying subject.  The ASTM subcommittee
D-22.11 dealt with this problem by writing the Standard Practice for
DETERMINING THE OPERATIONAL COMPARABILITY OF METEOROLOGICAL MEASUREMENTS -
D4430-84 (ASTM, 1984).  This work was patterned after Hoehne (1973) In which he
defines Functional Precision as the root-mean-square of a progression of
samples of the difference between simultaneous measurements made by identical
instruments collocated in the atmosphere.  Operational comparability applies to
two different kinds of Instruments rather than identical ones.   This Method
recognizes that, from an operational perspective, the precision of a
measurement can be estimated by know-ing how well Identical or similar
instruments measure the "same"  flow.

                     An EPA project collected data in Boulder,  Colorado  in 1982
to add to the literature some estimates of comparability. Finkelstein et al.
(1986) published in the refereed literature the material published by NOAA in
Kalmal et al. (1984).  Lockhart (1988) re-analyzed these data and concluded

-------
                                                             Section No. 4.2.2
                                                             Revision No.    Z
                                                             Date:   17 Sep 85
                                                             Page:   14  of 32


that the operational  precision of anemometers  at  10 m is no  larger than 0.2
m/s. Operational precision is the standard deviation of a series of difference
measurements which  is equivalent to the  operational comparability with all the
bias (mostly calibration error) removed.

                      An expression of accuracy for an anemometer operating on a
10 m tower  in  the atmosphere can be expressed  as  some function of speed  [f(u)l.
which comes from the  wind tunnel test, A(2), plus or minus 0.2 m/s.  This
estimate does  include the influence of turbulence on the sensor since the 0.2
m/s comes from collocated cups and vane  oriented  propellers  operating in a
turbulent summer environment.

                4.2.2.1.3  Distance Constant

                      4.2.2.1.3.1  Definition of distance constant

                      ASTM (1985a) defines distance constant  as the distance the
air flows past a rotating anemometer during  the time it takes the cup wheel or
propeller to  reach  (1-1/e) or 63 percent of  the equilibrium  speed after a step
change  in wind speed.  The step change is specified as one which  increases
instantaneously from 0 to the equilibrium speed.  The step change is simulated
by  releasing  a restrained anemometer in  a wind tunnel running at  the
equilibrium speed.   Several authors, among  them Acheson  (1988). Hayahsi  (1987).
Lockhart  (1987),   and Snow et al.  (1988), have commented on  the difference
between the distance constant to an increasing step function and  the distance
constant  to a decreasing step function.   The difference  is  larger with  larger
and heavier cup wheels, as is the size of the  resulting overspeeding error.
Snow et al. (1988)  point out that a system  including a sensor and an analog
signal  conditioner will have a combination  distance and  time constant.

                      4.2.2.1.3.2  Measurement  of distance constant

                      Most manufacturers will provide  the distance constant of
their  product.  These are usually derived from tests of  prototype sensors
during the  development phase of the product.   The variation from  individual  to
individual  in a production model Is not  large nor important.  It  Is  important
to  use a standard test and standard definitions  if distance constant
specifications are to be meaningfully compared to other  designs and
requirements.

                      EPA  (1987a) does not specify a distance constant  for
anemometers.   EPA  (1987b) does  suggest  in the  Instrument Procurement  section
8.1 a distance constant of  <5  m at 1.2 kg/m (standard sea-level density).   As
with accuracy, this  reference  uses a footnote  to specify the ASTM test method.

                      The  reason why distance constant  is Included is to urge
 users to buy  high quality responsive  sensors.   Heavy sensors with long distance
 constants  are more likely to produce  overspeeding errors,  which overstate the
 average wind  speed.   If  they are used to measure turbulence, they will
 underestimate sigma  u  because of a failure to  respond properly to eddy sizes
 smaller than  twice the distance constant.

-------
                                                             Section No. 4.2.2
                                                             Revision No.    G
                                                             Date:   17 Sep 89
                                                             Page:   IS  of 32
                4.2.2.1.4  Off-Axis Response
                This  specification, while included In ASTM  (1985) and
recognized in the  literature as a source of error, is not Included In EPA
requirements or suggestions.   It is mentioned here for completeness and  in
anticipation of future  specifications when more data have been published on the
subject.

                The off-axis errors from helicoid propellers are nearly  cosine
errors.  When a vane-oriented  propeller is turned in a wind tunnel so that the
wind is at some angle to  the axis of rotation of the propeller, the propeller
slows down.  The  Indicated speed from this misorientatlon of the propeller is
nearly equal to the total speed times the cosine of the angle of
mlsorientation.   That is, if the indicated speed from a propeller  is 5.00 m/s
and the propeller is  being held 10 degrees off  the true axis of the flow by the
aligning wind vane, the true speed is the indicated speed  (5) divided by the
cosine of 10 degrees  (0.9848)  or 5.08 m/s.  In  natural turbulent flow, a vane
located behind the propeller_may not keep the propeller perfectly  aligned with
the wind.  Small  misalignments result in small  errors since the cosine of a
small angle is nearly one.

                The off-axis errors from a cup  anemometer with a properly
oriented vertical axis  will depend on the design of the cup wheel  and  the angle
from horizontal from  which the wind reaches the cups.  MacCready  (1966)  and
Kondo et al. (1971) show  that  cup anemometers overstate the wind speed when the
air flow is not horizontal. Kondo shows the overestimation by  the cups  tested
to be 5 percent when  the  standard deviation of  the elevation angle is  17
degrees and 10 percent  at 25 degrees.  Siting on ridges or  building  tops or
anywhere the distortion of the flow over an object produces a  steady
non-horizontal flow will  result  in errors which will be unknown.

                The figure from MacCready  (1966) showing  the  response  of
various anemometers to  the elevation angle of  the wind  is  reproduced here as
Figure 4.2.2.9.
 -60  .50  '40  -30  -iQ  .10   0   10   20   30   40   50  60

              6 ELEVATION ANCLE (DECREES)
                                                       	(1) co. 0 CURVE
                                                       	|2) eo.» e CURVE

                                                       —— (3) STANDARD SMALL CUPS

                                                       ————(4) SENSITIVE SMALL CUPS

                                                       	 (5) TOTAL SPEED SENSOR

                                                       	(6) VELOCITYVANE
                                                       	(7) VECTORVANE
                                                         (AXIS CONSTRAINED HORIZONTAL)
Figure  4.2.2.9  Anemometer response to off-axis flow

-------
                                                             Section No. 4.2.2
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:   16  of 32
                 Consider the fact that a wind of  5 m/s with an elevation angle
of 30 degrees  will  have a horizontal component of 5 x cos  (30) = 4.33 m/s.  If
the presumption is  that the cup anemometer is providing the speed of the
horizontal  component of the wind, and if the cup performs  like a "total speed
sensor1'  in  the range of ±50 degrees as the figure suggests, the 5 m/s the cup
reports  is  a 15% overestimation of the true horizontal speed of 4.33 m/s.  A
propeller anemometer will report the horizontal component  because it does have
a nearly cosine response.  Operating side by side in this  30 degree wind, the
cup will report 5 m/s and the propeller will report 4.33 m/s and each will be
"right."

                 In  addition to the horizontal component dilemma, the cup
anemometer  tends to overestimate even the total wind.  This is particularly
noticeable  when the air is rising and flows past the support column creating a
wake which  interferes with the normal cup aerodynamics.  The figure shows this
effect  to be about  10% at +30 degrees for "standard small  cups."  This 30
degree  rising air example suggests that the side by side anemometers mentioned
above will  really be reporting 5.5 m/s (10% off-axis error for the cup) and
4.33 m/s for the true horizontal speed from the propeller, or a 27%
overestimation of the horizontal component by the cup anemometer.

-------
                                                             Section No. 4.2.2
                                                             Revision No.    0
                                                             Date:    17 Sep 89
                                                             Page:    17  of 32

       4.2.2.2  Wind Direction

                4.2.2.2.1  Threshold

                     4.2.2.2.1.1  Definition

                     As with wind speed measurement,  a key to a good wind vane
for air pollution applications is a low threshold.  The threshold is the one
performance characteristic which will certainly change with time because of
bearing degradation.  Most wind vanes use potentiometers to convert position to
output voltage.  Potentiometers have bearings or bushings which will wear and
add to the starting threshold.  There is not a standard definition for wind
vane threshold, although the ASTM Standard Test Method for DETERMINING THE
DYNAMIC PERFORMANCE OF A WIND VANE (ASTM. 1985b) offers the following
candidate.  The eg in the definition is the equilibrium direction of the vane
in a wind tunnel at about 10 m/s.

          Starting threshold (S , m/s) is the lowest speed at

          which a vane will turn to within 5° of 0  from an

          initial displacement of 10*.

Even this definition runs into some problems in interpretation.  If the vane
must move at least from 10° to 5° atothe threshold speed, is the offset
sensitivity really 5° rather than 10°?

                     The requirement in EPA (1987a) for PSD applications states
"Wind direction and wind speed systems should exhibit a starting threshold of
less than 0.5 meter per second (m/s) wind speed (at 10 degrees deflection for
direction vanes)."  Does this mean that a vane that moves from a 10°
displacement to 9.5  at 0.5 m/s has a starting threshold of 0.5 m/s?  The newer
EPA (1987b) on-site guidance says

          "Threshold (1)               s 0.5 m/s
          (1) as determined by wind tunnel tests conducted on production
          samples in accordance with ASTM D-22.11 test methods."

The reason the ASTM committee required the vane to move from 10° to 5° was to
relate the starting. threshold to accuracy.  With wind speed, there is a way to
correct for threshold nonllnearity; for wind direction there is not.  It seemed
best to establish the range of operating speeds to correspond to the range
where accuracy requirements are met.  ASTM assumed 5° for wind direction as a
reasonable accuracy.

                     When torque measurements began their use as a measure of
starting threshold, the question became clearer.  If the vane is required to
move to 5  there should be enough torque developed by the wind speed working on
the tail area exposed at 5° from the wind tunnel center line or the true wind
direction to turn the shaft assembly and transducer.   This sounds like a 5*
threshold requirement, and perhaps that is a better description.  As will be
shown later& there is a big difference between the torque developed at some
speed at 10  and the torque developed at the same speed at 5°.  The nature of a
standard test method Is less important than the application of a standard
method everyone uses and regulatory performance requirements consistent with
that test method.  This handbook will use the 10° offset moving to 5° on
release in the wind tunnel (the ASTM method) as the criteria for starting
threshold.  The relevant torque for this definition is that at 5°.

-------
                                                             Section No.  4.2.2
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:   18  of 32
                     4.2.2.2.1.2 Threshold measurement
                      The  measurement  of starting  threshold requires a wind
tunnel capable of accurate  operation  below 1  m/s.  One standard methodology is
defined in ASTM  (1985b) and described in Finkelsteln  (1981).  Just as it is
with wind speed, it  is possible to measure the torque which results from the
force of the wind on a wind vane as the torque measurement device holds the
vane at some angle from the wind tunnel centerline, say  10°.  Lockhart (1978)
describes wind tunnel test  data using two very different wind vane designs, the
front-damped Meteorology  Research. Inc. (MRI) Model 1074 and the more
traditional Teledyne Geotech (TG) Model 53.2.

                      Another body of  wind vane torque data exists as a result
of tests run by  the  R. M. Young Company (RMY).  Their tests used a DC "torque
motor" as the transducer  for vanes mounted in their wind tunnel.  The torque
motor current was  linearly  correlated to torque measured with a series of
Waters Torque Watches.  The torque motor drove the vane  to each of four
positions, plus  and  minus five degrees and plus and minus ten degrees from the
wind tunnel centerline.   A  measurement of current was taken at each position
and at each of 12  wind  speeds varying from 0.3 to 6 m/s, depending on the vane
design. Table 4.2.2.7 lists the average constant, K,  which was found by a
linear regression  of the  motor current (torque) to the square of the wind
speed, with the  intercept forced to zero.  They tested all of their products
along with some  vanes from  other manufacturers.

          Table 4.2.2.7 - K Values for Vanes  at Two Angles to the Wind
Vane Type
Offset Angle
5°
K
Wind Sentry (RMY 03301)
Wind Monitor (RMY 05103)
Wind Monitor AQ (RMY 05305)
Propvane (RMY 08003)
Mlcrovane (RMY 12302)
Blvane-19 cm fin (RMY 17003)
Anemometer Bivane (RMY 21003)
Propeller Vane-23 cm (RMY 35003)
Long Vane (Vaisala WAV 15)
Short Vane ' (Vaisala WAV 15)
Black Aluminum (Met One 024A)
High Damping Ratio (Met One024A)
F460 Vane (Climatronlcs 100075)
where: K = T/U2, E « Std. Err.
1
10
16
15
25
14
17
19
3
2
13
19
16
of
.8
.6
.8
.9
.0
.5
.1
.0
.6
.0
.8
.9
.0
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
Coeff.
E
006
080
126
061
414
188
141
127
047
015
181
194
322
and
8
10
O
K
3.
23.
37.
38.
57.
37.
45.
46.
7.
4.
28.

29.
r =
7
3
0
8
5
6
6
5
8
3
4

8
K(
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.

0.
10*

E
017
114
260
304
760
367
457
378
049
029
394

497
)/K(

r
2.
2.
2.
2.
2.
2.
2.
2.
2.
2.
2.

1.
5°)


1
2
2
4
3
6
7
4
2
2
1

9

                      The ratio of  the  10 degree K value to the 5 degree K value
 seems to be lower'for high aspect  ratio vanes.  A square vane has an aspect
 ratio of 1.  The Propvane, Mlcrovane,  Bivane and Anemometer Bivane are examples
 of designs with an aspect ratio  of 1.  A rectangular vane which is two times as
 high as it is long (along the tail boom) would have an aspect ratio of 2.  The
 Wind Monitors are examples of this design.  The F460 vane has an aspect ratio

-------
                                                           Section No.  4.2.2
                                                           Revision No.    0
                                                           Date:   17 Sep  89
                                                           Page:   19   of  32
of 4 and a K ratio of 1.9.  The "high aspect  ratio" TG Model 53.2.  whose  torque
data (natural log of torque vs. natural log of displacement angle)  are  shown in
Figure 4.2.2.10, has an aspect ratio of 6 and has "a K value ratio of  1.4  at
0.45 m/s.  Differences in torque between 5 degrees and 10 degrees could not  be
measured at 2.2 m/s  (the K. value ratio therefore equals one at that speed).
The High Damping Ratio (Met One) also has a high aspect ratio and also  could
not provide a stable torque reading at 10 degrees.    The MRI 1074  (aspect
ratio of 2) has a K ratio of 2.6.  This design is more difficult to compare  to
other vanes because of its front damping vane.

       Wind Vane  Torque  Data  -  Two Vanes
                     1975 Data - (see Lockhart 1978)
     8
                                                                     -500
                                                                     -100
                                                                         u
                                                                         I
              KEY:
           MRI Model 1074
           TG Model 53.2
                                                                      -50
                                                                         D
                                                                         cr
                                                        3.0
                     In Displacement Angle  (deg.)
Figure 4.2.2.10  Torque measurements as a function of vane angle.
                If  the starting torque of the shaft of a direction vane bearing
and transducer assembly Is to be interpreted in terms of wind speed, an
expression of torque as a function of speed is required.  Each expression is
specific to the vane design and an offset angle.   Take, for example, the Wind
Monitor AQ shown in Table 4.2.2.7.  The expression for a 10 degree offset Is

                              T = 37 U?

                If  a starting torque were found to be 5.9 g-cra,  that
measurement can be  expressed as a threshold wind speed of 0.4 m/s (0.9 nph).  A
0.41 m/s wind at a  10 degree angle from the vane position will produce enough
torque to move the  vane closer to the wind direction.  The expression for this
wind vane for a 5 degree offset is
                              T = 16.8 U?
The same starting torque of 5.9 g-cm will require a wind speed of 0.6 m/s  (1.3
mph) to move the vane  closer than 5 degrees to the true wind direction.

-------
                                                             Section No.  4.2.2
                                                             Revision No.     0
                                                             Date:    17 Sep 89
                                                             Page:    20  of 32
           Table 4.2.2.8 - Wind Vane Torque vs. Wind Speed and Angle
Wind
Speed
U
(m/s)
1.3
2.2
4.5
Offset Angle 0 (deg. )
5° A
T T:
(g-cm)(g-cm)
72 69
216 217
10° A
T T2
(g-cmj(g-cm)
72 76
166 177
562 557
15° A
T T3
(g-cm)Cg-cm)
108 97
324 278
1152 1163
20° A
T T4
(g-cm)(g-cm)
144 150
432 430
T is the measured torque holding the vane at offset angle 9
K is the linear regression coefficient when a = 0
T is the predicted torque using K from the following:
A i fi A 1 fi
Tj= 19.55 U T2= 50.22 U
A 2. A 2.
T3= 57.47 IT T4= 88.81 IT
                     The data from Lockhart  (1978) for the MR I Model 1074 are
shown in Table 4.2.2.8  to demonstrate  the complexity of the dynamic performance
of some vane designs.   A simple expression is useful to convert a torque
measurement to a wind speed.   The simple vane designs listed in Table 4.2.2.7
fit a U2 expression quite well.   The 5 degree and 10 degree data for the Model
1074  define a different slope than U   on the log-log plot of Figure 4.2.2.11.
An expression of U *6 fits the data well enough to use  to  extrapolate  the
experimental data for this vane design to other wind speeds.  The physical
reason for this unusual dynamic performance  is probably related to the effect
of the front damping vane and the relatively large support column.  The
vortices shed by the column only effect the  rear vane.

                     The question remains, should the 5 degree K value be used
or the 10 degree K value?  For the purpose of making a conservative estimate of
starting threshold for  performance accurate  to 5 degrees, the 5 degree K value
is recommended.  The user should not expect  this  torque-defined threshold to
agree with the "starting threshold" published by manufacturers.  Only after a
test is specified, like the ASTM test, can a 5 degree K value be expected to
agree with the data  sheet values.

                     4.2.2.2.1.3 Torque measurement

                     Starting torque measurements of a wind vane may be made  in
either of two general ways.   If the vane can be removed,  a torque watch can be
used to measure  the  starting torque of the bearing assembly and transducer (see
Figure 4.2.2.3).  For  this method to be most accurate, an equivalent weight of
the removed vane must  be placed on the shaft to simulate  the end loads of the
shaft of the bearings.

-------
                                                          Section No. 4.2.2
                                                          Revision No.    0
                                                          Date:   17 Sep 89
                                                          Page:   21  of 32
      8
         Wind  Vane  Torque  Data  - MRI  1074
                      1975 Data - (see  Lockhart, r978)
   E
   u
   I
   en

   (U
   3
   CJ
   O
   I—
   c
      7 -
6 -
5 -
      4 -
1.3  m/s     *  20°
            x  15°
                                2.2 m/s
                  KEY:
                A  10°

                    °
       T = 88.81 U2
                                                       4.5 m/s
                                                           i
                                       T  = 57.^7 U
                                                                      -1.00:
            T = 50.22 U1'6
                                   T = 19.55 U
                                                                 E
                                                                 G
                                                                 I
                                                                 cr
                                                                 cr
                                                                 x_
                                                                 O
                                                                |- 100
                                                                      - 50
        0.2      0.4      0.6       0.8       1
                          In  Wind  Speed  (m/s)
                                             1.2
                                                  1.4
1.6
Figure 4.2.2.11  Torque measurements as a function of wind speed.
                    If the vane cannot be removed or the choice is to not
remove it,  the starting torque can be measured by imposing a force at a
measured radial distance from the axis of rotation.  A spring-type gra« scale
at 10 cm from the axis of rotation will yield g-c» after dividing by 10.  On
some designs it is impossible to impose the force at 1 cm.   In the interest of
accuracy,  it is better to use a longer distance so the length part of the
measurement can easily be just a few percent. Of course the trade-off for
accurate distance is small force, an eo^ally troublesome source of uncertainty.
Figure 4.2.2.12 shows different gram scales used on a Young Wind Monitor AQ.
Figure 4.2.2.12   Starting  torque measurements on a vind vane.

-------
                                                             Section No. 4.2.2
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:   22  of 32


                      If the vane is left  on,  the space used for the measurement
must be devoid  of  any air movement.  Human breath provides a force which can
bias the measurement.   It is also iaportant that the axis of rotation be
vertical to negate any Imbalance in the vane  assembly.

                      For either method, the full 360 degrees of rotation of the
vane should be  challenged with the highest torque found being  reported as the
starting torque (worst case).

                 4.2.2.2.2 Accuracy

                      4.2.2.2.2.1 Accuracy definition

                      There is no transfer function  for a wind  vane comparable
to that for an  anemometer, unless the conversion of shaft position to output
voltage is taken to be such a function.   The vane is assumed to be accurately
placed, on average, downwind from the axis of rotation, when the  wind is steady
and its speed is well above the threshold.  If the  vane  is bent in some way, a
bias will be  introduced (see Dynamic Vane Bias later in  this section).  This is
seldom large  enough to be of concern.

                      The accuracy of the  sensor is  described by how well the
shaft position  is reported by the transducer and signal conditioning circuit.
The accuracy  of wind direction must include the accuracy with  which the sensor
is sited with respect to TRUE NORTH.  Any error in  orientation will be a true
bias and can  be removed from the data at  any time the facts become known.  The
"facts"  in  this case mean a rigorous quality control program which requires a
site log  to  indicate any servicing of the sensors.  The  "true  bias" can change
if the sensor is removed and reinstalled  without "as-found and as-left"
orientation measurements In the log book.  Any possible  undocumented change can
negate data correction for orientation.

                      The requirements for accuracy  Include EPA (1987a)  which
states  "Wind  direction system errors should not exceed 5 degrees. Including
sensor orientation errors."  In EPA  (1987b) it says
    "Accuracy (error) (1)      *3 degrees relative to  the sensor
                                mount or Index (s5 degrees
                                absolute error for installed  system)

      (1)  as determined by wind tunnel tests conducted on production
     samples  in accordance with ASTM D-22.11 test methods."

The footnote  is in error.  There  is nothing in the  wind  tunnel test which
relates  to wind direction accuracy.

                      4.2.2.2.2.2  Measurement of sensor accuracy

                      The  simple procedure for this  measurement requires soae
fixture which provides for  steps  in  the direction vane  shaft position of known
size.   There are  innumerable devices and  methods for this procedure,  many of
which  will be described  in  the calibration section (4.2.5.2).   One device Which
can move the shaft  in 60  degree  increments Is shown in  Figures 4.2.2.13.   The
 important criteria  are stability  and knowing that the error band for the
fixture is on  the  order  of  0.1 degrees of arc.

-------
                                                             Section No. 4.2.2
                                                             Revision No.    0
                                                             Date:    17 Sep 89
                                                             Page:    23  of 32
Figure 4.2.2.13
Wind direction calibration fixtures  from Teledyne
Geotech (left) and Met.  Standards Institute (right).
                     Typically, potentiometers used for wind direction will
have a linearity of about 0.5 percent, 1.8 degrees in 360 degrees.  A table of
angles and output values will usually fall within a range between -1.8 degrees
and +1.8 degrees when  the bias  is removed by subtracting the average error from
each error.  This statement  is  true when the open sector of the potentiometer
is ignored (for 360 degree mechanical and about 352 degree electrical systems,
see Section 8.2) or when errors in 540 degree format switching systems are not
considered. Other contributions to sensor error, such as hysteresis,
out-of-round and signal conditioning errors, when added to the linearity error
mentioned above should provide  an error band not larger than -3 to +3 degrees
relative, or 6 degrees if the bias has not been removed.

                     An example of audit data shown in Table 4.2.2.9 describes
the performance of one wind  vane when challenged with a 60 degree fixture.  The
fixture settings and the displayed digital output ofthe system are listed.  The
system had a 540 degree range and a 5 volt full scale output.  The output is
converted  to nominal voltage to show how the 540 range works.  (Degrees per
volt = 540 / 5 = 108)
                      The  average  error of -3.4
                               was calculated without using
                                 When the fixture is installed
the obvious "open  section"  values  marked by  '
the vane substitute  is  set  in the  180   location and then rotated until  the
output  is about  180.  This  need not be  precise since the average error  provides
a means of normalizing  the  data by removing  the initial bias of approximate
setting.  The  linearity of  the potentiometer-signal conditioner can be  seen  in
Figure  4.2.2.14.   Except for  the "open  sector" near 360 ,  the error is  within a
± 3   band, including the 540   format switching error of about 1 .

-------
                                                       Section  No. 4.2.2
                                                       Revision No.    0
                                                       Date:    17 Sep 89
                                                       Page:    24  of 32
     Table 4.2.2.9 - Relative Wind Direction vs.  Output Direction
CT>
a>
    1  -
IM
••5  C
i_
o -•
   -2 -
   -3
Fixture
Setting
A
(deg. )
180 ccw
120 ccw
060 ccw
360 ccw
300 ccw
240 ccw
180 ccw
120 ccw
180 cw
240 cw
300 cw
360 cw
060 cw
120 cw
180 cw
240 cw

cw
System
Output
B
(deg. )
177
114
054
001
298
238
176
115
177
239
292
001
056
115
177
239
average
Nominal
Voltage
B/108
(volts)
1.639
1.056
0.500
0.009
2.759
2.204
1.630
1.065
1.639
2.213
2.704
3.343
3.852
4.398
4.972
2.222
error C
is clockwise, ccw
Error

E=B-A
(deg. )
-3.0
-6.0
-6.0
1.0"
-2.0
-2.0
-4.0
-5.0
-3.0
-1.0
-2.0
1.0»
-4.0
-5.0
-3.0
-1.0
= -3.4 ("
Normalized
Error
E-C
(deg. )
0.4
-2.6
-2.6
4.4»
1.4
1.4
-0.6
-1.6
0.4
2.4
1.4
4.4*
-0.6
-1.6
0.4
2.4
values
Output
B-C
(deg. )
180.4
117.4
057.4
005.4
301.4
241.4
179.4
118.4
180.4
242.4
301.4
005.4
059.4
118.4
180.4
242.4
excluded)
is counterclockwise
         Relative  Wind Direction Accuracy
            Actual 16-Pomt Audit Results - 540  Degree Format
         Number indicates point sequence
          1—8  counterclockwise
          8-16 clockwise
                                    4 1 2
                                          * 10.16
                                 1X9.15
               +13
      /  +7
               + 3
+2
              060
                         300
                         120      180       240
                         Fixture Position (deg.)
Figure 4.2.2.14  Results of a wind vane audit using 60°  steps.
                                                           360

-------
                                                             Section No.  4.Z.2
                                                             Revision No.    I
                                                             Date:    17 Sep 8r
                                                             Page:    25  of 32


                     An analysis of this type helps to optimize the accuracy c:
the orientation.   If an orientation target is at 120°  TRUE,  when the vane is
pointing from 120° the output should read about 118°.   This effectively centers
the error band (see 4.2.4.3.2 on orientation).
                     4.2.2.2.2.3 Measurement of orientation accuracy

                     Orientation error is an important part of the measurement
error,  but it cannot be considered until the sensor is installed in the field.
The accuracy of the orientation includes the accuracy  in finding TRUE NORTH ana
the accuracy with which the vane is aligned to TRUE NORTH.   Use different
methods for finding TRUE NORTH.  Methodology for orientation is given in
4.2.4.3.2.

                     4.2.2.2.2.4 Expression of Accuracy

                     An accuracy specification should  include enough
information to define the type of accuracy intended and the method by which
accuracy claims may be tested.  There is no requirement for traceability to N2S
for wind direction.  The measurement of relative direction is a fundaacntal
division of a circle.  The measurement system can be bench tested by basic
methods yielding a clear expression of the errors associated with a position
angle vs. system output transfer function.  The errors are mostly conditional
biases which are small enough  (less than one percent of 360 degrees) to  ignore.

                     The orientation error is a pure bias which cannot be bench
tested.  The total error, a simple sum of the two parts (root-sum- square, RSS,
combination is only  legitimate with random errors, not biases), can only be
found after installation.

                     4.2.2.2.2.5  Precision
                     The definition of accuracy describes a bias term and a
variable  term akin to precision.  A comparability test (see 4.2.2.1.2.4) will
show that two vanes  properly sited and perfectly oriented will report the 20
minute scalar average directions with a difference of  less than two degrees,
i.e. precision is ±  2  .
                     An expression of accuracy for a wind vane operating on a
10 m tower in the atmosphere^can be expressed as the relative accuracy plus
orientation accuracy and ± 2*  for precision.  For a collocated test  (Lockhart.
1988), the orientation error can be estimated by the average difference  between
the subject wind vane and a collocated wind vane perfectly oriented.  If the
orientation error  is found to  be large, and if a quality control syste*  has
provided  records of  maintenance showing the orientation has not been changed, a
bias correction can  be applied.  The accuracy of the data corrected for  bias  is
then the  relative accuracy ± 2°.
                4.2.2.2.3  Delay Distance (Distance Constant)

                     4.2.2.2.3.1  Definition of delay distance

                     ASTM  (1985b) defines delay distance (D) as the distance
the air flows past a wind vane during the time  it  takes the vane to return  to
50 percent of the  initial displacement.   The value for this sensor
specification is  found  in wind tunnel tests, as described in Finkelstein
(1981).   The  initial displacement  is  10 degrees and D is the average of  a
series of tests at 5 m/s and  10 m/s using displacements on both sides of the
tunnel centerline.

-------
                                                             Section No. 4.2.2
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:   26  of 32


                      The specification in EPA (1987a, PSD) says "...the
distance constant  should not exceed 5  m."   In EPA  (1987b, On-Site) the
specification  says

     "Delay Distance (1)                 sS m at  1.2 kg/m3 (standard
                                           sea-level density)
       (1) as determined by wind tunnel tests conducted on production
         samples  in accordance with ASTM D-22.11  test methods."

                      4.2.2.2.3.2  Measurement of delay distance

                      Measurement requires a wind  tunnel of reasonable size and
quality.  The  width of the tunnel should be at least three quarters of  the
overall length of  the wind vane to be  tested.   With the small displacement
angle  of 10 degrees (about 3 percent of full scale), it is hard to conduct this
test in the open  atmosphere.

                      This specification is strictly a sensor dynamic
performance specification.  Any time constants in  the signal conditioning
circuits will  dampen the apparent sensor response  and make D larger than it is
for the sensor.   One could argue that  it is ONLY  the combination of D and the
time constant  of  the signal conditioner that should be considered in meeting
the regulatory requirements for performance. The  5 m maximum for D is roughly
equivalent to  a time constant of 0.5 seconds at 10 m/s wind speed and aim
vane.  For this and other reasons it is best to keep the time constant  of the
signal conditioning circuits to 100 ms or less.  For the same reason, it is
necessary to use  high speed recording  equipment for the wind tunnel tests.  At
10 m/s, aim  vane reaches the 50 percent D value  in 100 ms.  If one wants
resolution to  find D to 10 percent of  the true value (0.1 m), a 5 ms resolution
in the data  is desirable.

                4.2.2.2.4  Overshoot or Damping Ratio

                      4.2.2.2.4.1  Definition of overshoot or damping ratio

                      ASTM (1985b) defines Overshoot  (ft) as the ratio of the
amplitudes of  two successive deflections of a wind vane as it oscillates about
6  after release  from the offset position, as expressed by the equation
 B

                                    e
                                     '•*"
                                    e
                                     n
where e   and &      are the amplitudes of the n and (n+1)  deflections,
       n       f n * 1 t
respectively.

The  Damping Ratio (TJ) may be calculated approximately from the overshoot  ratio
by the formula
                                    "•
                                         -

-------
                                                             Section No. 4.2.2
                                                             Revision No.    0
                                                             Date:    17 Sep s9
                                                             Page:    27  of 32

                  The specification in EPA (1987a,  PSD)  says "The damping
ratio of the wind vane should be between 0.4 and 0.65..."    In EPA (1987b,
On-Site) the specification says

              "Damping Ratio (1)        £0.4 at 1.2 kg/m3   or

               Overshoot (1)            <25'/. at 1.2 kg/m3

               (1) as determined by wind tunnel tests conducted on
               production samples in accordance with ASTM  D-22.11 test
               methods."

                     The subject of dynamic wind vane performance is thoroughly
discussed in MacCready and Jex (1964), Gill (1967), Weiringa (1967) and Acheson
(1970).
                     4.2.2.2.4.2  Measurement of overshoot

                     The measurement of overshoot also requires a good wind
tunnel and sensitive, fast response recording systems.   A  series of tests were
conducted by Lockhart in 1986 in pursuit of a wind vane design with a 0.6
damping ratio.  A sketch of the results of this unpublished work is shown in
Figure 4.2.2.15 to provide an example of how various vane  designs compare in
overshoot and delay distance.   One of the requirements in  the ASTM method is an
initial offset of 10 deg.

                     4.2.2.2.5  Dynamic Vane Bias

                     The Dynamic Vane Bias (6 , deg. ) Is the displacement of
                                             8
the vane from the wind tunnel centerline at 5 m/s.   This measurement will
Identify wind vanes with unbalanced aerodynamic response because of damage
(bent tail) or design.  This is a screening specification  not needed or used in
any application requirements.     The ASTM method measures  this difference, if
any, and disqualifies the vane if the difference is greater than one degree.

       4.2.2.3  TURBULENCE

                4.2.2.3.1 Definition

                The Glossary of Meteorology (Huschke, 1970)  quotes Sutton
(1955) defining turbulence as a state of fluid flow "in which the instantaneous
velocities exhibit irregular and apparently random fluctuations so that in
practice only statistical properties can be recognized and subjected to
analysis.  The situation is, in fact, analogous to that accepted unreservedly
in the field of molecular physics..."  The definition is ended with a quote
from the Bible"

               The wind bloweth  where it  listeth and  thou  hearest
               the sound thereof but canst not tell whence it cometh
               and whither it goeth:...  John 3:8

                From the standpoint of wind measurement, then, turbulence is
not measured, it  is calculated.   From the standpoint of quality assurance,
turbulence is a difficult subject to control.  It is possible to define the
measurement samples from which the statistical properties are calculated.  It
is possible to define the algorithm by which the samples are summarized.  The
relationship between the algorithm and the application or model is also
important, but it is beyond the scope of this handbook.

-------
                                                                 Section No  4.2 2
                                                                 Revision No.    C
                                                                 Date:    17 Sep 89
                                                                 Page:    28  of 32
       •  Delay Distance  (50* recovery)
                   Thies (experimental)
                   697 grams
                           MRI 1074
                           656 grams
                           Thies
                           600 grams
                    Climatronics F460
                    251 grams
                    MSI (experimental)
                    191 grams
                    Climet
                    170 grams
                   Vaisala
                     92 grams
                    MSI  (experimental)
                     72  grams
   012        5

  Distance  (m) at 5 m/s                              Meteorological Standards
                                                     Institute - March 5.  1986
Figure  4.2.2.15  A  sample of  the dynamic response of some wind  vanes

-------
                                                             Section No. 4.2.2
                                                             Revision No.    C
                                                             Date:   17 Sep 89
                                                             Page:   29  of 32


                4.2.2.3.2 Direction Measurement for Sigma Theta

                The most common turbulence property routinely reported is signs
theta, the standard deviation of a series of horizontal wind direction samples.
Most, but not all, of the following will also apply to Sigma phi. the standard
deviation of a series of vertical wind direction samples.   Among the
specifications which are important to the direction measurement used to
calculate sigma theta is delay distance which limits at the small end of the
eddy size spectrum the eddy sizes to which the vane can react.  If the vane has
a delay distance of 5 m, it will not detect energy from eddys smaller than 5 s
because the vane cannot react to them.  If 1 m eddy sizes are important to the
diffusion being estimated, use a wind vane having a delay distance of 1 m or
less.
                Another important specification is overshoot or damping ratio.
Vanes will overshoot when correcting for a direction change. If the overshoot
ratio is 0.5 (or 50%), more variability will be reported from the same
turbulent flow than is reported by a vane with an overshoot ratio of O.25  (or
25%).  The relationship between overshoot ratio and damping ratio is given in
Table 4.2.2.10 as calculated by the equation found in 4.2.2.2.4.1.

          Table 4.2.2.10 - Overshoot Ratio vs. Damping Ratio
Overshoot
Ratio
1.00
0.90
0.80
0.70
0.60
0.50
0.40
0.35
0.30
0.25
0.20
0.15
0.10
0.05

Percent
100
90
80
70
60
50
40
35
30
25
20
15
10
5
! denotes
Damping
Ratio
0.00
0.03
0.07
0.11
0.16
0.22
0.28
0.32
0.36
0.40
0.46
0.52
0.59
0.69
PSD range
                Sigma calculations are biased by any averaging built into  the
signal conditioner.  They are also subject to error if external noise gets into
the output, a dilemma for circuit designers.   A compromise might be to filter
out any noise at frequencies higher than 20 Hz (0.5 m at 10 m/s).  In winds
above 20 m/s, this filter would degrade data from a wind vane having a delay
distance of 1 m.           Turbulence from mounting structures upwind of the
vane will bias  the sigma value.  Out-of-balance conditions with a vane
measuring sigma phi will also bias the statistic, particularly at the low  wind
speeds.  Un-filtered noise from potentiometers will add an error to the natural

-------
                                                             Section No. 4.2.2
                                                             Revision No.    C
                                                             Date:   17 Sep 89
                                                             Page:   30  of 32

variability of the wind.  These are problems  which are best detected by
inspection of strip charts or  oscilloscope  traces.

                4.2.2.3.3 Statistical  Summaries

A few basic concepts will help in considering the specifications of the
statistical algorithm used and the representativeness of the value calculated.
Here again, careful definition will help understand what the circuits and
logical networks are doing to  the input  samples.  EPA (1987b) devotes 44 pages
to Meteorological Data Processing Methods.

                Representativeness is  the important concept to keep in mind
when examining strange or unusual data.   Samples of wind direction taken over a
short period of time  (seconds  to a minute or  two) are likely to exhibit nearly
normal or Gaussian distribution.   As the time gets longer  (a few minutes to an
hour or more), physical dynamics driving the  flow in the surface layer may
provide different shapes.  The most common of these might  be the bi-modal
distribution resulting from  land-water,  mountain-valley, day-night or
meso-scale convectlve flow systems. Whatever the driving  forces, a bi-modal
distribution cannot be usefully represented by a mean and  standard deviation.
This is to say that a data sampling and  processing system  may work perfectly
and produce numbers which have no physical  meaning. From a specification
standpoint, tests for "working perfectly" are possible and should be used.

                The method used by the wind direction system to describe the
position of the vane  in the  series to  be statistically described must be
thoroughly described and understood.  The most common error in the past,
perhaps even made today, is  to do nothing.   If the output  voltage unambiguously
represents an azimuth angle, and if samples of voltage are described with the
statistical parameters of mean and standard deviation, and then expressed in
units of azimuth angle, great  errors will result.  These errors are a result of
a discontinuous range of output voltage.  If  001-360 degrees are represented by
0-1 volt, samples clustered  around 360 will contain some near 0 and some near
1.  The mean of 0.5 will be  180 degrees  away  from the mode.

                When analog  ink recorders were used exclusively with 360 degree
formats, it was common  to see  the paper  painted  red by the pen going back and
forth through full scale, effectively obliterating any data.  There are several
ways to avoid or minimize the  "crossover" problem.  System specification should
define how this will be done.   The most  common method for  minimizing this error
is to use a "540 degree" format.   Systems were designed with dual
potentiometers or dual wipers  180 degrees out of phase.  When the wiper moved
into the gap, circuit switching would change  to  the center of the other circle.
This switching would be  invisible in the output  at  the 1/3 and 2/3 scale
points, but when the voltage went beyond full scale, it would switch to  1/3
scale and when the voltage went to zero, It would switch to 2/3 scale.  This
format completely eliminated the pen painting problem and  drastically reduced
the output voltage switching,  but some large  pulses remained to occasionally
bias sigma calculations.

                With  the advent of microprocessors and digital computers  it
became possible to combine  the samples without any  large pulses.  One method
uses a unit vector sum  to find the resultant  vector direction  (average
direction).  With an  assumed wind speed  of 1  m/s, each sample of wind direction

-------
                                                             Section No. 4.2.2
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:   31  of 32

is converted from polar coordinates  (001-360 degrees] to cartesian coordinates
(N-S and E-W components in meters). The components are added or subtracted over
the sample period and  the resultant vector direction is found by  converting the
final coordinate sums  to polar coordinates through an arc-tangent calculation.
The standard deviation of each sample about the resultant vector  direction is a
straight forward process on the differences of the samples from the "average."
Of course, this same method can be used with the true speed-weighted wind
vector samples.

                Some automatic systems currently available pick an assumed
direction, usually  the mean of the last period, take the digital  difference in
degrees (limited to 180) of each sample from the assumed mean, and find  the
standard deviation  and the mean difference for the period.  The standard
deviation about the mean is the same as the standard deviation about the mean
plus a constant.  The  mean difference plus the assumed direction  is the  true
mean (limited  to 360 degrees).

                There  are also other algorithms which estimate the standard
deviation (see Turner, 1986).  It  Is only necessary from the standpoint  of
quality assurance to know that the method used is being satisfied with the
samples taken  from  the measurement system.

                The sample size is specified In EPA  (1987b) as 360 samples to
estimate  the standard  deviation to within 5 or 1054.  Lockhart  (1988) found an
apparent bias, not  a random error, when the standard deviation was estimated
from 120  samples over  a 20 minute  period.

                Most models accept data representing one hour.  Sigma theta for
60 minutes is  Influenced by the changing wind direction during the hour.   It  is
recommended  (EPA, 1987b) that four 15-minute sigma theta calculations be
combined  to provide a  "one hour" value for the purpose of selecting a
Pasquill-Glfford stability class.  The method is
2 + 2 + 2 + 2 '
A1S *30 *45 *60
4
1
2
                  A(l-hr)


              where 
-------
                                                              Section No.  4.2.2
                                                              Revision No.    0
                                                              Date:   17  Sep 89
                                                              Page:   32   of 32
is excluded  from this value but included in the true standard deviation  of  the
direction  about  the hourly mean.  If the standard deviation about  the  hourly
mean Is  required for a concentration distribution analysis,  the correct
formulation  for  a-     .  is shown below.
                  A(l-hr)
 A(l-hr)
                                BI + n2 + n3 + n4
                                                                     I
                                                                     2
  where n ,n ,n ,n  are the number of samples In periods 1,2,3 and

     and d2,d2,d2,d2 are calculated for each period from
                             -  Vs
                          d3-  VA
                          d4a   VA
              where A   is  the  average direction  for period n^ etc.

                and A   is  the  average direction  for all four periods.

 See Box et  al.  (1978)  for further discussion of the  calculation of  total
 variance  from discrete subset variances and means.

-------
                                                             Section No. 4.2.3
                                                             Revision No.    0
                                                             Date-   17 Sep 89
                                                             Page:    1  of  5
4.2.3  ACCEPTANCE TESTING
The procurement document, purchase order or contract, should be specific at
least in terms of required performance specifications.  "Required" in  this
context may only mean that the instrument meets the suggested or specified
regulatory performance.  It is another question, beyond the scope of this
handbook, whether "necessary" relates to the application to which the  data will
be used.

    There are two kinds of performance specifications, those which can be
verified by simple inspection testing and those which require unusual  test
equipment and experience.  The former should be tested and the latter  certified
by the manufacturer.  The manufacturer should have either performed these tests
on one or more samples of the model design or arranged for such tests  to have
been conducted by some calibration facility.  In either case, a test report
should be available to any who require the documentation.  The cost of the copy
of the test report should not be much larger than the cost of normal copying.
If the manufacturer does not provide such documentation, the choice is between
accepting the manufacturer's unsubstantiated claim or having a specific test
run.

    A good QA Plan will provide for a QA sign-off of  the procurement document
in order to assure that equipment capable of the required performance  is being
purchased and that the capability can be verified.  Purchasing by brand name is
often expedient where the performance of a model has  been verified and all that
is required is more units.  This practice Is also cost effective when
considering spare parts and instrument technician training costs.

    The parts of wind speed and wind direction  sensors which predictably
deteriorate and seriously Influence the performance of the sensors are the
bearings.  It is acknowledged that an experienced inspector can "measure"
bearing condition by feeling or spinning the shaft.   The receiving inspection
is a protection against putting defective equipment  into the field.  It  is not
a necessary link in the documentation trail for data  validity purposes.  True
torque measurements for data validity will be most valuable at the initial
field calibration.  True laboratory conditions  may be chosen, however, because
torque measurements are sufficiently difficult  to make.Therefore, a receiving
inspection may be used for this purpose.

       4.2.3.1 Wind Speed

       An example is found in Table 4.2.3.1.  This performance specification
for an anemometer is hypothetical but one which will  meet the requirements of
EPA  (1987a. PSD).  Each attribute of the instrument  is Identified by a key as
to whether it is a receiving test candidate of  not, and  the nature of  the
testing  is briefly discussed below.  Each instrument  includes a sensor,  signal
conditioner, and recorder.  When an attribute of  the  sensor is affected  by the
signal conditioner or  recorder, a keyed comment will  be made.

-------
                                                             Section  No.  4.2.3
                                                             Revision No.     0
                                                             Date:    17 Sep 89
                                                             Page:     2  of  5
             Table 4.2.3.1  -  Anemometer Performance Specification
            Range

            Thresholdlt2(t)
                            1.2,
            Accuracy (error)  '  (t)(*)
                              2
            Distance Constant (t)
  0.5 to 50 m/s

s 0.5 m\s

a (0.2 m/s + 5% of observed)
                 3
s 5 m at 1.2 kg/m  (standard
  sea level density)
               (t) as determined by wind tunnel tests conducted
                   on production samples in accordance with
                   ASTM D-22.11 test  methods  (ASTM, 1985a).
               (t) aerodynamic shape  (cup or  propeller) with
                   serial number to be  accompanied by test report,
                   traceable  to NBS.  showing  rate of rotation vs.
                   wind speed at 10 speeds with 0.1 m/s resolution.

                1  subject  to receiving inspection
                2  transducer with signal conditioner
                4.2.3.1.1   Threshold

                The  threshold receiving test  should examine the system output
with the anemometer  not  turning (below sensor threshold) and with the
anemometer turning at  an equivalent 0.5 m/s.   If the cup is a Climatronics F460
Vinyl Cup Set  (100083) or  Heavy Duty Cup Set  (101287), the published constant
for mph and 30 pulses  per  revolution is 9.511 which converts to 1.41 meters per
revolution.  At 0.5  m/s  the cups should be turning at

                             0.5+1.41  =0.35 rps,

or one revolution  in 2.8 s.  The cup assembly can be turned by hand to
approximate that rate  of rotation.  If the anemometer  is a Young propeller
(08234, 18 cm  polypropylene), the turning factor is 0.294 meters per
revolution.  At 0.5  m/s  the propeller should be turning at

                             0.5 + 0.294 =1.70 rps,

or two revolutions in  1.2 s.  This also can be approximated by  turning  the
propeller by hand.

                The  key measurement for threshold, however,  is  starting torque.
This  requires  knowledge of the K value  (see 2.1.1.2) which should be available
from  the  manufacturer.

                4.2.3.1.2  Accuracy

                 The receiving test for  accuracy is the conversion of rate of
rotation  to output in units  of wind speed.  The transfer  function,  supplied by
the manufacturer,  should be  in terms  of rate of rotation  (rps)  vs.  wind speed
 (nips).  The receiving inspector simply  turns the anemometer  shaft at a  few
known rates of rotation to  see if  the systemoutput compared  to the  predicted
output is within the tolerance specification.

-------
                                                             Section No. 4.2.3
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    3  of  5
                4.2.3.1.3 Distance constant

                The distance constant determination requires a special wind
tunnel test and is beyond normal receiving inspection capability.  The time
constant of the anemometer circuitry will influence the effective system
performance.  Assume the manufacturer's value for distance constant of the
sensor is one meter.  At a wind speed of 10 m/s the sensor will have a time
constant of 0.1 s.  Assume the time constant of the system electronics is 2 s.
Then, at a wind speed of 10 m/s, the system time constant is 2 s or 20 m.  In
this example, the system electronics would need a time constant of 0.1 s or
less if the response capability of the sensor is to be fully available.  At a
wind speed of 0.5 m/s the sensor responds in 2 s, the same as the system
electronics.  In this example, the system electronics dominates the sensor
response at all speeds.  If the sensor response is to be available at all
speeds (up to 25 m/s), the electronics time constant must be 0.04 s or less.
The time constant can be measured at the receiving test by timing how long it
takes for the output to reach 63.2% (1-1/e) of a step change in speed.   The
step change can made by turning the anemometer shaft at a known rate of
rotation and then instantaneously stopping its rotation.

       4.2.3.2  Wind Direction

       An example of a wind direction specification is found in Table 4.2.3.2
(see EPA, 1987b).  This performance specification is also hypothetical but it
is one which will meet the requirements of EPA (1987a, PSD).  Each instrument
includes a sensor, signal conditioner, and recorder.  When an attribute of the
sensor is affected by the signal conditioner or recorder, a keyed comment will
be made.

              Table 4.2.3.1 - Wind Vane Performance Specification
              Range

              Threshold1>2(t)
              Accuracy (error) '  (t)
              Delay Distance (t)
              Damping Ratio (t)
              Overshoot2(t)
                          001 to 360 degrees or
                          001 to 540 degrees
                        a 0.5 m\s

                        s 3 degrees relative to the
                            sensor mount  or index -
                        s 5 degrees relative to
                            TRUE NORTH

                        a 5 m at 1.2 kg/m3 (standard
                            sea level density)

                        a 0.4 at 1.2 kg/m3
                        s 25% at 1.2 kg/m
(t)  as determined by wind tunnel tests  conducted on
    production samples in accordance with
    ASTM D-22.11  test methods (ASTM,  1985a).
 1   subject  to receiving Inspection
 2   transducer with signal conditioner

-------
                                                             Section No. 4.2.3
                                                             Revision No.    0
                                                             Date:    17 Sep 89
                                                             Page:     4  of  5
                4.2.3.2.1   Threshold
                The  threshold receiving test  is  a  starting torque measurement
(see 4.2.2.2.1).  To relate  the torque measured  to wind speed and off-set
angle, a K value is  required,  either from the manufacturer or from an
independent test.  The  torque measurement may be made with the vane assembly
removed or with the  vane assembly in place.   If  the  latter is chosen,
verticality is essential to  negate any out-of-balance in the vane assembly from
biasing the test.  Also there must be no air  motion.  Very small air motions
will bias the test.   Use a smoke puff to be sure the air is still and refrain
from breathing in the direction of the vane surface.

                4.2.3.2.2 Accuracy

                The  receiving Inspection is the  best place to establish the
true non-linearity,  if  any,  of the direction  vane  transducer.  A test using
some circle dividing fixture capable of fine  resolution, 1 deg. for example,
will provide a record which can be referenced in future field spot checks.
Without such a test  It  is hard to prove accuracy of  s 3 deg.  If several units
show the same pattern of non-linearity, it should  be acceptable to sample
future units and accept a generic shape of the error.  When a long series of
samples is planned,  there is a tendency to devise  methods which are quick.  The
time constant of the signal  conditioning circuit must be known to establish the
minimum time between position change and output  reading or recording.   If the
step is small,  like  1 deg..  three time constants will deliver 95% of 1 deg.,
which is good enough.  If the step is large,  like  180 deg., three time
constants will deliver  95% of 180 or 171 deg. which  is not good enough.   It
takes seven time constants to deliver 99.9% or 179.8 deg.

                The  receiving Inspection cannot  include the orientation error.
The manufacturer does not deliver orientation.  There may be orientation
fixtures, however, which assume that an optical  center line  is parallel  to the
line set by an  orientation pin.  This assumption can be tested.  Field
orientation may be based on the orientation of a crossarm with the assumption
that  the output angle when the vane is parallel  to the crossarm is known.  This
assumption can  be  tested or the alignment fixture set in  laboratory  conditions
to  the desired  output.

                4.2.3.2.3  Delay distance and overshoot

                These dynamic characteristics require a special wind tunnel
test  and  their  determination is beyond normal receiving Inspection capability.
The time  constant  of the wind direction circuitry can influence the  system
performance as  it  can with wind speed.  Assume the manufacturer's value for
delay distance  is  one meter and the time constant of the  electronics is 2 s.
At  wind  speeds  of  10 m/s  the time constant of the wind vane  is 0.1 s and  at 0.5
m/s It  is  2 s,  the same as the system  electronics.  For this example.
therefore,  at  all  speeds  above 0.5 m/s the performance of  the wind vane is
being limited  by the time constant of  the electronics. The  time constant  can  be
measured at  the receiving test by timing how  long It takes  for  the output to
reach 63.2% (1-1/e)   of  a  step  change  In direction.  For example,  the step
change  can made by quickly turning  the vane from 000 to  180 deg.  The  time
constant is the time required  for the  system  output to  change  from 000 deg.  to
 113.8 deg.

-------
                                                             Section No. 4.2.2
                                                             Revision No.    I
                                                             Date:   17 Sep c?
                                                             Page:    5  of  5
       4.2.3.3  Measurement System
       All the elements of a system of signal conditioners, recorders and
monitors will require checking for correct function.  The receiving inspectic-
should Include testing these various sub-systems.  There may be a calibration
switch which replaces the sensors with simulated conditions.  Assume a systen
has a calibration switch which substitutes the equivalent of 25 m/s and  180
deg. to the input of the signal conditioning boards.  In 4.2.3.1 and 4.2.3.2
the sensors were providing the input values.  It is possible for these tests to
show perfect results and yet the outputs with the calibration switches on could
show the system to be out of calibration.  The reason would be that the
adjustments for the substitute calibration inputs were off.

       After the calibration inputs have been adjusted and the "output"  shows
the system to be in calibration, a parallel analog recorder may show incorrect
values.  This event could be caused by an incorrect adjustment in the interface
which drives the analog recorders from the output.  So far the "output"  is
assumed to mean the voltage which goes into the data logger and becomes  the
archived data.   There may be monitoring meters or digital displays on the
system panel.  These monitoring meters may differ from the "output" because
they have individual adjustments.  All the sub-systems should tell the same
story and the receiving inspection should verify that they do.  In fact. It is
rare when a system arrives in receiving with the various outputs in
disagreement, but they must be checked.

       It goes without saying that the receiving function records the model
numbers and serial numbers of the component parts and checks the parts received
against the purchase document and the shipping document.

-------
                                                             Section No. 4 2.4
                                                             Revision No.    Z
                                                             Date:    17 Sep 59
                                                             Page:     1  of  5

4.2.4  INSTALLATION

       4.2.4.1  General Considerations

       From a QA point of view, there are aspects of the installation which
should be considered.  Perhaps the most important of these is siting.  See
4.0.4.4 for general siting criteria and discussion.   From a QA point of view,
however,  failure to meet the recommended siting criteria may be necessary.  If
the general site is selected for other measurements for good reason, the wind
sensor siting may be only a best compromise.  There are considerations which
set the options for the compromise.  Among these are technical and budgetary
considerations.  The qualitative Judgments which go Into siting are briefly
discussed here.

       If buildings or trees are likely to interfere with the wind speed or
direction sensing, try to locate the tower or pole such that the wind sensors
will most faithfully record speed and direction for the direction of primary
concern,  e.g. for directions that would take an effluent toward a residential
area.
       Another important technical consideration is accessibility of the
sensors.   There should be no hesitation in taking a hands-on look at the
sensors whenever a performance question arises.  Yet this Is the most difficult
task at most sites.  Some sites require special "climbers" to retrieve a sensor
and wait to return it to its installed position.  These field people may not be
trained to handle delicate instruments.  It may be so difficult or expensive to
get a sensor down that suspected bad data will be accepted rather than facing
the problem.  This reluctance causes mounting hardware to become corroded to
the point that the sensor cannot be removed without damage.   Most of the
breakage of delicate sensor parts results from handling while climbing on a
tower.  If the direction sensor does not have an alignment fixture, it may not
be possible to remove the sensor without going through the orientation
procedure at re-installation.

There are several ways to overcome most of these problems.  They all require
design forethought in Installation.  First, the sensors need to be easily
removed from the mounting structure.  They need to be as easily connected to
the rest of the measurement system when they are down, either with the same
cable or a suitable substitute cable.  One popular cup and vane design uses a
crossarm which contains non-removable cabling.  Either the whole crossarm
assembly must be taken down with the sensors or there has to be a substitute
crossarm to plug the sensors into at ground level.  Some towers, the short 10 •
types, can be tilted to access the sensors while still connected.   In this case
the sensors are about 90 deg. from their operating position.   Some tests
require the sensors to be vertical.  In these cases the sensor still needs to
be removed and re-connected.  Ideally, the crossarm is left exactly as is so
when the sensor is replaced, nothing physical has changed with respect to
vertlcality or orientation.   Some towers telescope for access to the top where
wind sensors usually are mounted.  This makes access easier but removal and
reconnection is still necessary.  Some towers have elevators which transport
the sensors up and down the tower.  How they deal with cables depends on the
elevator design.

       An ideal installation is one where the operator can get hands on the
sensor, perform a test adjacent to the electronics and recorder, and re-install

-------
                                                             Section No. 4.2.4
                                                             Revision. No.    G
                                                             Date:   17 Sep £9
                                                             Page:    2  of  S

the sensor,  safely,  alone and within one hour.   There are  no technological
reasons why  this  cannot be done, except for tall towers where an elevator may
take 20 minutes for  a one-way trip.   There probably  is not a budgetary reason
for avoiding something like this ideal installation, once  the cost of invalid
data and  true operating costs are factored into the  formula.  It is usually not
done simply  because  the need for service is overlooked and the method of access
is not pre-planned.

       4.2.4.2  Wind Speed

       The wind speed sensor is most susceptible to  error  from shadowing and
interference.  Aside from the need to have the anemometer  properly exposed, the
only other consideration is verticallty (for cup anemometers).  If the cup
wheel is  well balanced, a small angle (1 deg.  or less) in  mounting is not
important.   If the cup wheel is not well balanced,  the starting threshold will
be degraded.

       4.2.4.3  Wind Direction

                4.2.4.3.1  Exposure

                The  problem with vertical!ty for the direction vane is Just the
same as with wind speed.  For a well balanced vane  assembly, a small angle from
vertical  is  not  important.  If the vane assembly is  not well balanced, the
starting  threshold  is raised and a predominant direction for light winds may
not have  any basis  in fact.

                Wind vanes are often used for dispersion applications by
calculating  the standard deviation of the wind direction about the mean
direction, sigma  theta.  Unless the wind vane is at  the tower top, there will
be some direction where the wind goes through and around the tower before  it
gets to the  vane.  The farther from the tower the wind vane is mounted the
smaller the  sector with tower interference.  The interference sector can be
selected  by  placing  the vane on the appropriate side of the tower.

                4.2.4.3.2  Orientation

                Of  all the sources of error for a wind direction measurement,
the orientation of  the vane to TRUE NORTH has the potential and reputation of
being the largest.   A bad orientation provides a fixed bias to the data which
can be  removed.   If  the vane is moved and the constancy of the bad orientation
is in question,  the  data may not be recoverable. The method of wind vane
orientation  must  be  capable of 1 deg. accuracy with 2 or 3 degrees as the upper
limit of  the error.   Two steps are necessary to achieve an oriented wind vane.
First,  the location of TRUE NORTH must be found to  an accuracy of less than 1
deg.  Secondly,  the  wind vane "index" must be aimed at that location with  an
accuracy  of  better  than 2 deg.  (see 4.2.2.2.2.3 for a discussion of the
location  of  the  normalized error "index").

                 TRUE NORTH as distinguished from magnetic  north is usually
found by  reading  a  magnetic compass and applying the correction for magnetic
declination.  The declination can be read from a USGS map.  The Fox Island
Station declination, according to the 1959 (revised in 1981) map, is 20.5°.
The USGS  is  now  providing a computer service called GEOMAG. See below:

-------
                                                             Section No. 4.2.4
                                                             Revision No.    '.
                                                             Date:   17 Sep Er
                                                             Page:    3  of  =
CONNECT
        Unauthorized use of this U.S. Government computer system
                     Is punishable under PL98-473
Welcome to the USGS Branch of Global Seismology and Geomagnetism
               On-line Information System
Type Q for Quick Epicenter Determinations (QED)
     H for Historical Epicenter File Searches  (EIS)
     M for Geomagnetic Field Values
Enter program option (Q, H or M): m
                                  GEOMAG
The International Geomagnetic Reference Field  (IGRF) was revised in August,
1987.   The models for 1945.0, 1950.0, 1955.0, and 1960.0 have been superseded
by new definitive models (see, for example,  EOS Transactions, American
Geophysical Union, vol. 69, no. 17, April 26, 1988, pages 557-558). The new
models were installed on June 21, 1988.  Please note that the revision affects
field values for dates between 1945.0 and 1965.0, but not those for later
dates.
Problems or suggestions?  Please contact Norman Peddle,  U.S. Geological Survey.
MS-968, Federal Center, Box 25046, Denver. CO 80225. Telephone:  (303) 236-1364
(FTS 776-1364).
Press RETURN to continue:
Do you want information about this program (Y/N)?      [ 1  N
Options:   1) Field Values (D, I, H, X, Y. Z. F)
           2) Magnetic Pole Positions
           3) Dipole Axis and Magnitude
           4) Magnetic Center                          [1]   [ ]
Display values twice (Y/N)?                            [Y]   [ ]
   Name of field model:                    [         ]          ?
The following field models are available:
     Name             Type               Date range
                                       1945.0 - 1990.0
                                       1985.0 - 1990.0
                                       1985.0 - 1990.0
                                       1985.0 - 1990.0
   IGRF85      Spherical Harmonic
   USCON85     Spherical Harmonic
   USALA85     Spherical Harmonic
   USHAW85     Spherical Harmonic
Name of field model:                {         ]
   Date:                             1/25/89]  [
Latitude:
North or South (N/S):                 ]  N
Longitude:
East or West (E/W):                   ]  W
Elevation:                          [0.0]  [250 feet
                                                    ]47<
      Region
    World
    48-States
    Alaska
    Hawaii
 USCON85
  1
.25
                                                    ]122.6292
Model: USCON85    Latitude  : 47.25    N
Date : 1/25/89    Longitude: 122.6292 W
                                               Elevation:   250.000 ft
D
deg mln
19 47.9
19 47.9
I
deg mln
69 29.0
69 29.0
H
nT
19526
19526
X
nT
18372
18372
Y
nT
6613
6613
Z
nT
52181
52181
F
nT
55715
55715
Annual change:
    0  -5.5    0  -0.7    -1.9
    0-5.5    0  -0.7    -1.9
                                   8.7   -29.8   -39.6   -37.7

-------
                                                             Section No.  4.2.4
                                                             Rtvision No.    I
                                                             Date:   17 Sep Er
                                                             Page:    4   of  5


                The  GEOMAG program was accessed by calling  1-800-358-2663
through  the modem of an "XT clone" using a "ONE TO ONE"  communications program.
Using the capture feature of ONE TO ONE, the following  communication was
recorded.  Note that the Fox Island Station declination for 1/25/89 is  19.8  cr
0.7° less than the map indicated.  This is consistent with  the  0 deg. -5.5 mir..
annual change for the roughly eight years since the map was revised.

                The  other way to find the direction to  TRUE NORTH employs sone
astronomical observation.  While the compass method is  clearly  easiest,  it  is
also the most prone  to error.  Good training and equipment  will reduce  these
errors to an acceptable level, but not the "less then  1 deg." advised above.
Training will minimize errors from the influence of nearby  metal objects  and
the mis-application  of the declination correction, but  local variation  in the
isogonic field is unknown.  On the other hand, the observation  of astronomic
bodies can be unambiguous.  Polaris, the north star, will provide TRUE  NORTH  to
within 1 deg.  (without correction) on any clear night.   The true  solar  noon
method will provide  the north-south direction to within 0.1 degree  on any clear
day, given  the station longitude, date and an accurate clock.   A  simple Basic
program  will provide the azimuth angle  to the sun at any time of  day given the
station  longitude,  latitude and date.  Examples of the two  solar  methods are
given below.

                      4:2.4.3.2.1 True Solar Noon Method

                      The True Solar Noon  (TSN) method  finds the time at some
particular  date  at  some particular  longitude when the  sun is in the north-south
plane passing  through the North  Pole, the South Pole and the longitude
selected.   If  the sun is not directly overhead  (elevation 90 deg. )  the  azimuth
line  to  the sun is  TRUE SOUTH or  TRUE NORTH.  Two calculations are required.
First,  find the time of  the  Local Apparent Noon (LAN)  from the longitude.  The
examples shown here are  for:
             Fox Island. WA  (Long.  122.6292, Lat.  47.2500).  07/04/90  and
             New Or leans, LA  (Long.   90.1100, Lat.  30.0000)  12/25/90.
T   = 12:00:00 + 4(Long.  - 15n),  where  n is  the number  of  time zones from
  LAN
Greenwich.   Table 4.2.4.1  is a  list of  n values for United States time zones.

                             Table 4.2.4.1 Time  Zones
Time Zone
Eastern
Central
Mountain
Pacific
Yukon/Alaska
Hawaii
n
5
6
7
8
9
10
 T   (WA)= 12:00:00 + 4(122.6292 -  [15 x 8]) = 12:10.52 = 12:10:31 PST
  LAN
 T   (LA) = 12:00:00  +  4(090.1100  - [15 x  6])  =  12:00.44 = 12:00:26 CST
  LAN

 Secondly,  correct  for  the Ephemeris of the sun.
 T    = T    - A, where A is the correction found in Table 4.2.4.2.
  TSN    LAN

-------
                                                             Section  No.  4.2.4
                                                             Revision No.     C
                                                             Date:    17 Sep 89
                                                             Page:     5  of  8
                      Table 4.2.4.2  Ephemeris of  the Sun
From the Nautical Almanac - 1 989 Yachtsman's Edition
Equation
Date of time
m. s.
Jan. 1 - 3 24
4 -448
7 -608
10 -724
13 -835
16 -940
19 -1039
22 -11 31
25 -12 16
26 -12 54
31 -13 25
Feb. 3 -13 49
6 -1405
9 -1414
12 -1416
15 -14 11
18 -1400
21 -13 42
24 -13 18
27 -1249
Mar. 2 -12 15
5 -11 37
8 -10 55
11 -1010
14 -921
17 -831
20 - 7 38
23 -644
26 - 5 49
29 -454

Equation
Date of time
m. s.
Apr. 1 - 4 00
4 -307
7 -216
10 -126
13 - 39
16 + 6
19 +48
22 +1 26
25 +159
28 + 2 29
May 1 +2 53
4 +313
7 +328
10 +338
13 +342
16 +342
19 +337
22 +326
25 +311
28 + 2 52
31 + 2 28
Jun. 3 + 2 00
6 +129
9 +55
12 + 19
15 - 19
18 - 58
21 - 1 36
24 -215
27 -253
30 - 3 30
Equation
Date of time
m. s.
Jul. 3 - 4 05
6 -437
9 -506
12 -532
15 -553
18 -609
21 - 6 20
24 - 6 27
27 - 6 28
30 -625
Aug. 2 -615
5 -601
8 -540
11 -515
14 -444
17 -408
20 -328
23 -243
26 - 1 55
29 -103
Sep. 1 - 8
4 +50
7 +150
10 +252
13 + 3 55
16 +500
19 +604
22 +708
25 +811
28 +912

Equation
Date of time
m. s.
Oct. 1 +10 11
4 +1106
7 +1202
10 +1252
13 +1338
16 +1420
19 +1456
22 +15 27
25 +15 52
28 +1610
31 +16 21
Nov. 3 +1625
6 +1622
9 +1612
12 +15 54
15 +15 28
16 +14 55
21 +14 14
24 +13 26
27 +12 31
30 +11 29
Dec. 3 +10 21
6 +908
9 +751
12 +629
15 + 5 04
18 + 3 37
21 + 2 08
24 + 38
27 - 51
30 - 2 20
   TT«.(WA>  •  12:10:31  - (-4:16)
    T5N
   TrsH(LA)
12:00:26 - (-0:08)
12:14:47 PST

12:00:34 CST
                                    13:14:47 PDT
                Once the time of TSN is known, all that remains is to observe
the position of the Line to the sun at TSN.  An easy way is to use a loosely
mounted theodolite set at 180 deg. to track the sun.  [CAUTION: EYE DAMAGE MAY
RESULT FROM LOOKING AT THE SUN WITHOUT SUITABLE PROTECTION - Remember that
harmful UV rays can be present when visual light "looks" safe - USE AN EYE
SAFE FILTER]  When a watch (one second resolution set to WWV or equivalent
source for the correct time) shows TSN, tighten the theodolite mounting.  At

-------
                                                             Section No. 4.2.4
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    6  of  8


that Instant the  sun is in the cross hair of  the  theodolite and the theodolite
correctly labels  the azimuth angle as 180.  Once  set, the theodolite can be
used to find the  bearing to any distant feature which might be selected as an
orientation target.   Another method is to nark the  end of the shadow of a
vertical tower at TSN,  thus establishing a N-S line from the base of the tower
to the mark.

                      The two drawbacks to the TSN method are weather and
schedule.  If the sun Is obscured at TSN the  observation cannot be made.  Also,
if other activities  command higher priorities,  the  time of TSN might not be
available for the sighting.

                      4.2.4.3.2.2 Solar Azimuth Method

                      The azimuth angle to the sun can be found at any time if
the latitude is also known.   A Basic program  (Blackadar, 1985) which contains
the necessary subroutines has been edited to  provide the outputs shown  in
Figure 4.2.4.1.   These are the same two examples  as are used in 4.2.4.3.2.1.
The program listing  Is given as Figure 4.2.4.2.  Notice in Table 4.2.4.3 that
the two methods do not agree.  The differences are  trivial.  Even at the fast
angular motion of July, the sun moves about 0.5 deg. per minute.  The roughly
quarter minute difference in methods represents only a little over 0.1  deg.
uncertainty.  Notice also the nonlinear!ty difference between winter and summer
which makes simple extrapolation impossible.

                      A Brunton compass, mounted on  a tripod, can be used for
solar sighting.   The mirror can be set to project the sun and the sighting
points and lines  on  a white piece of paper.  The  compass needle can be  used as
a reading index or an additional protractor and pointer can be added to the
compass mounting  hardware.

-------
                                                                    Section  No. 4.2.4
                                                                    Revision No.     0
                                                                    Date:    17 Sep 89
                                                                    Page:     7  of  8
DAY? 4
MONTH? 7
YEAR? 1990
   DAP 25
   ffflmP 12
   YEARi 1990
                      SOLAR AZIMUTH ANGLE
                      HEDNEEDAY 4 JUL 1990
                Fox Island. HA
                Longitude 122.6292   Latitude 47.25
                Day of Year IBS      Julian Day 2448077
                  Transits Meridian  13  14  S3 PDT
                     SLH AZ1KITH ANGLE
                     TUESDAY  25 DEC 1990
                ten Orleans, LA
                Longitude 90.11     Latitude 30
                Day'of Year 39     Julian Day 24
                  Transits Meridian 12  0 34 CT
Tiie
HR,MIN? 11.30

HR,HIN? 12,00

HR.HIN? 12,30

HR.HIN? 13,00

HR,MIN? 13,30

HR,MIN? 13.00

Elevation

S7.81

61.36

64.00

6S.43

6S.42

6S.42
Azuuth

130. IS

141.90

155.86

171.73

IBB. 40

171.73
TlK
M.IIIK? 1C.30

K.H1K? 11.00

ffi.KIN? T..30

«SfNK? 12.00

H»,NIK? 12.30

ffi,MIN? 13.00

Elevation

^.35

34.67

34.11

36.61

36.14

34.73
A:iwth

155.23

1*3.07

171.33

179.86

138.36

196.65
Figure 4.2.4.1  Screen printouts for two Azimuth examples
                  Table 4.2.4.3   Solar Method Comparisons
         Lo.La.D          UA (7/4/90)
         Method     TSN      Almanac  dif.
         Units      PDT        PDT    (s)
         TSN    13:14:47  13:14:53  -6
         LA (12/25/90)

   TSN      Almanac  dif.

   CST        CST     (s)

12:00:34  12:00:34   0
              where  Lo  is the station  longitude
                      La  is the station  latitude
                      D   is the date of  interest

-------
   SO STAJ.-MS1. Fox Wind. WA-jREM STATION NAME^V "•
   '60 READ LO.IA
   100 DATA 122.629,47JSflEMLONGmjOE 4 LATmjOE~"  ~*
   110 READ PI.OB.LO.L1.AO.A1.EC.EO"
   120 DATA 3.141S92654..4090e5.4.88376619I.Ol7202791
   130 DATA6.23471229..017201970_016728..002ie
   140 TR-PV1 BOfC-Z-PlflEU TO RADIANS: FULL CIRCLE
   ISO SL-1S'INT(LO/15+.5):REM STANDARD LONGITUDE
   180 TZ-Sl/15-4: REM SELECTS TIME ZONE LABEL
   100 LO-LO-TRiA-LATRSL-SLTR
   210D1S--SUNOAY  MONDAY TUESDAY WEDNESDAY*
   220 D2S-THURSDAY FRIDAY SATURDAY*
   230 DS-01V»D2S: XS-* ~ •
   240 MS-*JANFEBMARAPRMAYJl*NJULAUGSEPOCTNOVDEC*
   250 Z$»*ASTESTCSTMSTPSTYSTASTAOTEDTCDTMDTPDTYDTADT*
   260 TN-LO/FC+.5:REM LONGITUDE TIME OFFSET * 12 HH
   270 PRINT "DAYTINPUT D
   280 PRINT •MONTH'jdNPUT U
   290 IF M>12 THEN PRINT 'INVALID OATE*«!OTO 280
   300 PRINT -YEAR-pINPyT YR
   310 X-1 :Y.1 .-QOSUB 2410
   320 JI-TflEM YEAR DAY 1
   330 X-D:Y-M:GOSUB 2410
   340 YO-T-J1 »1 AEM DAY OF YEAR
   350 X-INTfrviy7:Y-INT(X)
   360 WD-INT+ J)fl£M DAY OF WEEK
   380 T-T+3449.54TN:REM T S NOW TIME OF LOCAL MEAN NOON
   385 DT-.00059+2JE-08T: T.T+OT: REM EPHEMERIS TIME
   390 PRINT TAB(28)rSOLAR AZIMUTH ANGLE*
   40S PRINT TAB(2B):
   410 PRINT MID$(D$.9-WO+1.9}:
   420 PRINT D:MID$(MJ,3-(M-1H.1.3)iYR
   423 PRINT TAB(20):STA$;
   426 PRINT TAB(SO)rU>n9Jtud»Ufym;TAB<42)rUftu«i«-a-WTR
   430 PRINT TAB(20).-D«y of Y««-^rDiTAB(42)rJU!cn D«y^NT(JO*1)
   490 X-YD-WD: Y^L-ISTR
   500 IF X>90 AND X<29B THEN TZ-TZt-7 : SL-Y
   510 T$-MID$(Z$.3TZ+U)
   610 GOSUB 2860: REM FIND SUN AT LOCAL MEAN NOON
   620 IF DE>PI THEN  DE-DE-FC
   630 O-ML-RAflEM EQUATION OF TIME (NOT DISPLAYED]
   640 D3.OE: REM SAVE DECL FOR HEAT BUDGET
   680 X-.0145439 : GOSUB 2360
   690 IF ABS(Y)<1 THEN 720
   710 GOTO 780
   720 SO-2-(1+L1/FC): H^SO : GOSUB 2260
   725 TC..00274'SO-SINtOB)'COSfrU-SlN(LA)
   730 Z^IN(SO)'COS(l>)'(COSpE)*3)
   735TC-TOZ
   740 X-ZT+TC*EO: GOSUB 2310
   780 PRINT TA8(23).Tranilts MMUtn 1
   790 IF ABS(LA-DE>PI/2 THEN PRINT XSXSfTS
   800 H-0: GOSUB 2260
   810 X-ZT: GOSUB 2310
   820 PRINT X;Y2:TS
   830 PRINT
   850 PRINT TAB(12)rnmt      EMvnon  AamuDf
   1090 GOTO 4000
   20100-0*4^)
   2020 F YoO THEN 2050
   2030 Z-0£.1:IF X<0 THEN N-1
   2040 GOTO 2060
2osoz>x/r
                                                                                     Section No.  4.2.4
                                                                                     Revision  No.       0
                                                                                     Date:     17 Sep 89
                                                                                     Page:       8   of   8
2070 F C-l THEN Z-PV2-2
2060 F N.I THEN 2-2
2090 FYCOS(LA)>COS{DE)*COS(H)
2165 SZ^ORII-CZ^): ZAWkTN(SZ/CZ)
2170 F ZAUM.r(T-TN+(SL*ZVFC) :NEXT U
2280 ZT^: RETURN
2310 FX<0 THEN X-X*FC :GOTO 2310
2315 W.JT24/FC : X.INT(W)
2320 Z-COS(DE})
2370 f ABS1 THEN 2390
2380X^QRn-Y*2) : GOSUB 2010
2390 RETURN
2410T^6T{YR-1980)
2420 T.T-«T(r(YR*INT((Y««yi 2))M)
2430 S^SGN(Y4)^UABS«A1 T: REM SUN'S MEAN ANOMALY
2870 ML-UM.1T: REM SUffS MEAN CELESTIAL LONGfTUDE
2880 X^«(UL):Y.COS(UUa3OSUB 201 0
2890 Mt^Z
2900 DL*2*EC-SN(MAH-25*EC'2>SIN(2>MA}
2910 TAUIA*01_TL-ML*OLflEM TRUE ANOMALY A LONGITUDc
2920 RVK1-€C*2)/(1+ECPI THEN Z>Z-FC
 2960 RA-ZAEM SUN'S RIGHT ASCENSION
 2970 RETURN
 401 0 ZTH4«*MIN/60 : ZT-ZTH*FC/24
 4020 T.T-TN*
-------
                                                             Section No.  4.2.5
                                                             Revision No.    0
                                                             Date:    17 Sep 89
                                                             Page:     1  of  4
4.2.5  CALIBRATION
    Calibration,  as defined on page 3 of the Purpose statement in the beginning
of this handbook,  qualifies the process as both a measurement and adjustment,
if necessary, of  the performance of the system and its components.
Manufacturers usually include in their manuals the details of all the available
calibration or adjustment points.  From a QA standpoint,  the important
consideration is  how the system is working as a whole.   Since only parts of the
system are adjustable, the relationship of these adjustments to the whole
system must be known.  This brief section will focus on documentation of
calibrations and methods to verify the system response to subcomponent
adjustments.

       4.2.5.1  Wind Speed

                4.2.5.1.1  System accuracy

                The part of a calibration which challenges the entire system,
except for the coupling or reaction of the sensor to the wind,relates the rate
of rotation of the anemometer shaft to output speed.  It does not matter if the
rate of rotation is caused by a synchronous motor or a d.c. motor with a
provision for shaft revolution counting.  What does matter is the accuracy of
the determination of AVERAGE rate of rotation and a common averaging PERIOD
used by the system and the challenge.  The operators may choose to conduct this
calibration with the sensor Installed on the tower.  When multiple outputs
exist, the calibration should record values from each of them, but the critical
output is the one used to produce the official archieved data.


                The accuracy determination depends on both the method used in
the challenge and the accuracy of the measurement of the input.  If a
synchronous motor is used, there must be some reason to believe the motor was
turning in sync with the commercial power.  Repeated samples which do not
change is one form of evidence.  Commercial power is generated within a
frequency tolerance of 60 ± 0.1 cps.  Synchronous motors which are hand held
with a flexible coupling to the anemometer shaft may go in and out of sync
providing a slightly changing output.  Shaft rotation counters can also produce
erroneous outputs.  Some evidence of their performance, such as counting a
synchronous motor shaft rotation or simply counting revolutions at a slow rate.
is needed in the documentation of the test equipment, preferably before and
after field use.
                4.2.5.1.2 Component accuracy

                If the system has built in calibration circuits, they should be
calibrated at the same time as the total system.  They are handy to use on a
routine service schedule, but there needs to be some evidence of their
calibration.  If panel meters or portable OVOMs are used to check the signal
conditioner or transducer sub-system, there needs to be evidence that they are
in calibration.  It  is possible  to adjust a circuit to provide the required
output on a meter which has a 2Y. error and thereby  introduce a 2% error to the
calibrated system output.

                The  calibration of the sensor starting threshold can only be a
measurement.  Adjustment is usually  Impossible.  The exception might be the
amount of end play in the shaft-bearing assembly, but that level of sensor

-------
                                                              Section  No.  4.2.5
                                                              Revision No.    :
                                                              Date:    17 Sep 59
                                                              Page:     2   of  4

repair  is  usually left to the laboratory or shop for good reason.  The accuracy
of the  torque  measurement, or non-measurement,  is also important.  Assume a
torque  watch,  or similar device, with a range of 0.003 to 0.030 oz-in.  The
threshold  of measurement is 0.003 oz-in or 0.22 g-cm.   If a cup anemometer has
a K value  of 1.4 (see 2.1.1.2),  the torque provided  by a  0.4 m/s  (0.9 mph) wind
is 0.22 g-cm  [from T=Ku2].  The torque provided by a 0.5  m/s (1.1 mph) wind is
0.35 g-cm.  If the torque watch cannot measure the starting torque of the shaft
because it turns before the indicator moves,  the starting torque  is < 0.22 g-cn
and the starting speed is < 0.4 m/s.   If,  instead, the starting torque reads
0.35 g-cm  (0.005 oz-in), the starting speed is 0.5 m/s and within
specification.   If the starting torque reads 1.0 g-cm (0.014 oz-in or about
half scale on  the torque watch), the starting speed  is 1.4 m/s and clearly in
need of service.   Documentation of this measurement  will  tell the data QC
inspector  that the data from this anemometer is in error  In the indicated 0.2
(assuming  a 0.2 m/s offset) to about 3 m/s-range.  (3 m/s wind provides 12.6
g-cm torque or about an order of magnitude more than that provided at 1 m/s.
The difference between 0.35 and 1.4 is not likely to be visible at 12.6)  The
true wind  speed will oe higher tnan the indicated speed.   At indicated 0.3 m/s
it would be 1.5 m/s and at indicated 3 m/s it would  really be 3 m/s.

        4.2.5.2  Wind Direction

                 4.2.5.2.1 System accuracy

                 The system calibration of a wind vane can be checked  on the
tower by aiming the vane to and from known directions,  such as a distant
mountain peak  or similar feature.   If checks are made with respect to a mounted
component,  such as a crossarm, the orientation of the crossarm also needs to be
checked.   A single distant feature should be the orientation target with  a
known bearing  with respect to TRUE NORTH.   Other targets  can be secondary
checks  which challenge both the orientation and the  performance of the
transducer.  For systems using the 540 format,  the targets should be  reached
after a clockwise revolution and then again after a  counter clockwise
revolution to  challenge both parts of the transducer.

                 Before the transducer is removed from the tower, a
documentation  of the as-found output with the vane held pointing at or from the
orientation target is essential.  This single act provides the basis  for  data
validity for the period beginning with the previous  as-left record and ending
with this  as-found reading.  Since the sensor should not  have been removed from
the tower  or adjusted in orientation without the as-found and as-left readings.
these values should be the same, within the capability of pointing the vane (1
deg. , 2 deg. at the most).  If they are not,  the data QC  inspector will have
some detective work to do.  Usually,  when the sensor is removed and used  in a
calibration at the location of the rest of the system,  replacement in a keyed
fixture will cause the as-left value to be the same  as the as-found.   Figure
4.2.5.1 shows  three examples of how manufacturers provide an orientation  key
for wind direction sensors.  If there is no keyed fixture,  the full orientation
procedure  will be required.

-------
                                                              Section No.  4.2.5
                                                              Revision No.     0
                                                              Date:    17 Sep 89
                                                              Page:     3  of  4
Figure 4.2.5.1  Methods  for keying  orientation  of direction  sensors

                One  simple orientation procedure requires a  clamp which  will
hold  the vane from turning.  A hose clamp will  work for some designs.  Figure
4.2.5.2 shows a hose clamp used for this
purpose.  Tape which does not stretch is
marginally useful.   Stretchy tape like
duct  tape or electricians tape will only
work  on a perfectly  calm day.  Set  the
vane  so that the output  is the correct
value for the orientation target  (see
4.2.2.2.2.1).  If the angle of the
orientation target is coincident with a +2
deg.  error relative  to the average  error
of 0  deg., the output should be 2 deg.
higher than the bearing  of the orientation
target.  Only in this way will the  relative
error of the sensor  be distributed  equally
about TRUE directions.    Tighten the clamp so the output is both correct  and
constant.  Mount the clamped sensor on the tower and turn it until the vane
points at the orientation target.    Clamp the vane in place.   Verify that the
output is still correct  before removing the vane clamp.

                4.2.5.2.2 Component  accuracy

                The same comments regarding calibration circuits,  parallel
recorders and panel meters apply to wind direction as they do to wind speed,
mentioned in 4.2.5.1.2 above.   With  the sensor next to the signal  conditioner
(attached with either the operating  cable or a suitable substitute)  and with a
fixture which holds known relative directions,  the signal conditioner can be
adjusted if required.  The 540 offset voltage,  if one is used,  can be tested
and adjusted.   The output voltage vs. position can be set.   The open space in
the potentiometer,  if one is used,  can be measured and adjusted for.
    Figure  4.2.5.2
A direction vane clamp

-------
                                                             Section No. 4 2 5
                                                             Revision No.    I
                                                             Date:   17 Sep Sr
                                                             Page:    4  of  4


                 A  single potentiometer has  an electrical range of something
like 355 deg. with a mechanical range of 360 deg.   If the  transfer function of
relative direction to voltage output is
                                   9 = 360 x V

where G is  the  angle in deg.  and V is the output  in volts  (0-1 V scale), and
the maximum "full  scale" output, set by shorting  the potentiometer wiper to the
high side of the potentiometer, is 1.000 V,  a small error  will have been set
into the system.   The error will be +1.4 %  of reading.  At 355 deg. the output
will be 360.  At 180 deg.  the output will be 182.5  deg.  This adjustment error
added to the linearity error of the potentiometer may be more than is
acceptable.   If Instead, the signal conditioner is  set  to  output 0.986 V when
the vane is set to 355 deg.,  the output will be 355 deg. (360 x 0.986).  At 180
deg. the output will be 180 (assuming no linearity  error).   All of the error
between 355 deg. and 360 deg. is in that 5  deg. sector.

                 Is this acceptable for PSD  (EPA,  1987a) applications?  The
"wind direction system error" which cannot  exceed 5 deg. is  the error of the
averaged wind direction samples.  If the mean direction were 355 deg. with a
range of ±5 deg. ,  and if the distribution were bi-modal with half  the values at
350 deg. and half  at 360 deg., and if the output  voltage remained  at 0.986 V
between 355 deg. and 360 deg. , the average  output would be 352.5 deg. , a -2.5
deg. error.   If the dead space were at 0 V,  the output  would cause the bi-modal
distribution to look like half 350 deg. and half 360 deg.  producing the correct
average of  355  deg.  This is a maximum error estimate.  True distributions
would cause smaller average errors.  Even a wind averaging 357.5 deg. with a
range of ±2.5 deg  (the vane is always in the dead space),  the error is 2.5 deg.

                 The starting threshold of the wind  vane is important  to
accurate  low wind  speed directions.  The design of  the  vane  along  with  the
off-set angle (or  error tolerance) provides a K value.  The  K value along with
the starting torque of the vane assembly provides a threshold wind speed.
Assume a 5  deg. error tolerance and a K value of 15.  At 0.5 m/s  the available
torque is  3.75  g-cm.  At 10 cm out from the axis or rotation, a force of 1/3 of
a gram should move the vane assembly.  This is another  threshold of the  torque
gauge situation.  At 1 m/s the  torque available is  15 g-cm and at  10  en  the
force is a  reasonable 1.5 g.  At 1 m/s and 10 deg.  error tolerance, K becomes
37.5, the  torque available becomes 37.5 g-cm, and the force  at  10  cm  is an
easily measured 3.75 g.

-------
                                                             Section No.  4.2.6
                                                             Revision No.     0
                                                             Date:    17 Sep 89
                                                             Page:     1  of  2


4.2.6  OPERATIONS,  MAINTENANCE AND QUALITY CONTROL

       4.2.6.1  Operations

       The important aspects of operations, from the standpoint of quality
assurance, are planning (see QA Plan,  Section No. 4.1.)  and documentation
(Section No.  4.9.1).  The purpose of operations is to acquire valid data.   For
wind measurements,  this requires frequent (weekly, if possible) visual
examination of the sensors.  This is not a "hands-on" examination but simply a
look at the active shapes, cups, propellers and vanes, to be sure there has
been no physical damage.  Sensitive wind instruments can be damaged by hail and
by birds.   The nature of an analog recording, if one is  used routinely,  will
tell how the sensor is performing.  Routine entries in the station log will
provide the evidence of attention to support validity claims.

       Calibrations are a part of operations.  A member  of the operating
organization needs to become the "expert" on how the measurement system works
and what it needs to continue "in control" performance.   Regularly scheduled
calibrations build the expertise and the documentation showing measurement
accuracy.   The frequency of calibrations is a variable.   For a new
installation, a calibration during the installation is necessary.  A careful
look at the first week of operation will find early failures.  If all seems to
be going well, a calibration check after a month is prudent.  If no problems
surface, a full calibration at the end of the first quarter is advisable.  For
some site environments and some applications quarterly calibrations are
recommended.   Semi-annual calibration is the minimum frequency. If problems are
found they must be documented and corrected as quickly as possible.  The
requirement of 90% Joint frequency of valid wind and stability data does not
permit much down time.  The frequency of calibrations or calibration checks
should be determined by the performance of the instrument system.  If problems
occur, the week-month-quarter frequency should begin again.  When it is
demonstrated that the system is once again "in control," routine calibration
frequency (semi-annual or quarterly) can resume.

       4.2.6.2  Maintenance

                4.2.6.2.1 Routine and preventive maintenance

                The only routine maintenance required for the wind system
should be applied during routine calibrations.  Sensors exposed to the elements
need cleaning and protective lubricants applied to their mounting hardware.
When a sensor needs to be removed for close  inspection or calibration and  it
cannot easily be removed because set screws or nuts are locked to their threads
by corrosion, a failure in routine maintenance is the reason.

                If the system has supply requirements, such as ink and paper
for analog recorders or tapes and printer paper for digital recorders, the
timely servicing of these requirements is a  routine maintenance task.

                Preventive maintenance must at minimum follow the
manufacturer's recommendations.  Considerable damage can result by ignoring
this guidance.  Some people like  to oil anything  that moves.  Sensitive wind
sensors require specific care if  the threshold is to be maintained.

-------
                                                             Section No. 4.2.6
                                                             Revision No.    :
                                                             Date:   17 Sep S?
                                                             Page:    2  of  2
                4.2.6.2.2  Corrective  maintenance
                Parts for wind systems  are  not  to be found anywhere except fron
the manufacturer.   This is true at least  at the sub-component and component
level.  When a part fails or wears out,  the new part usually must come from the
manufacturer.  This may take a week or  two  depending on the part and the
manufacturer.  It  is prudent to have spare  parts on hand to cover some
predictable failures.   A component plug-in  philosophy is the quickest way to
correct failures.   If a bearing or a potentiometer fails in a sensor, a new
calibrated sensor  is simply plugged in  while the failed one is repaired.  If a
circuit card fails,  a new calibrated card is plugged in while the failed one is
repaired.

                The next level of spare part strategy is the sub-component
level.  Critical and difficult to buy parts are stocked and used to repair
sensors or circuit cards.   Conventional sensors will always need repair at soae
point in  time, bearings and direction potentiometers usually.  Circuit cards
are becoming so reliable that maintenance is hard to anticipate.

       4.2.6.3  Quality Control

       The quality control (QC) of a data monitoring program is a loop driven
by routine inspection of the data for validity.  The data QC person should be a
meteorologist who  is familiar with how  wind data should look and what kinds of
variety are provided by the atmosphere.   Such an Inspector will spot problems
before they are obvious to an observer  who  may  be an expert in another field
but is not a meteorologist.  If a technically qualified QC inspector is not
available, the best compromise that is  available must be made.  It is very
dangerous In terms of lost time if no routine data QC function is followed in
the QA Plan, or if there is no QA Plan.

       When a problem is found by the data  QC inspector, a discrepancy report
is issued which brings the operators into the data QC loop.  Their inspection
and corrective action is reported back  to the QC inspector closing the loop.
Because of this QC loop, the measurement system can be operated "in control"
and valid data produced.

-------
                                                             Section No  4.2.7
                                                             Revision No.    I
                                                             Date:    17 Sep 89
                                                             Page:     1  of 2C


4.2.7  PERFORMANCE AUDIT METHODS

       4.2.7.1  General Considerations

       A performance audit is the determination of the instrument system
accuracy made with an independently selected method and by a person who  is
independent of the operating organization.   To make this determination for wind
measurements, knowledge of the input conditions imposed upon the sensors is
required.  Given knowledge of these input conditions,  the transfer functions
and the system's data handling method, the output can be predicted.  The
difference between the predicted output and the system output is the error of
the system or its accuracy.

       The methodology starts with the ways of controlling and/or measuring the
input conditions.  When controlled inputs are used, as should always be  the
case for starting thresholds, anemometer rate of rotation vs. output and
relative vane position vs. output, the accuracy of the output is easily
determined.  Of course, the accuracy of the anemometer transfer function is not
a part of this determination.  When the input conditions are not controlled,
as with the collocated transfer standard (CTS) method, the accuracy
determination has a larger uncertainty.  The CTS method does challenge the
anemometer transfer function.  The best performance audit uses both methods
where appropriate.

       A performance audit must follow some written procedures.  Since the
procedures must be relevant to the design of the instrument or system being
audited, only general principles will be described below with some specific
examples.  The data from  the audit should essentially fill out an audit  form.
It is important, however, for the auditor to be sufficiently experienced to be
able to deviate from the  procedure or the form when the pursuit of truth leads
away from the expected.

       The starting point of an audit form is the documentation of the who,
what, where, when, and how the audit values were acquired.

                4.2.7.1.1  Who

                The performance audit report form should contain a space to
identify the auditor.  The audit report which summarizes the audit findings
should report the names and affiliations of the operators of the system.

                4.2.7.1.2 What

                The form  should contain a section to identify the  instrument
being audited by manufacturer, model number, and serial number.
Sub-assemblies, such as a cup wheel of an anemometer, should be identified by
number.  If  they are not  numbered, the operator should be asked to mark  them
for  identification.

                The audit report should contain a list of all the equipment
provided and used by the  auditor, including model and serial numbers and time
of last calibration, where relevant.

-------
                                                             Section No. 427
                                                             Revision No.    0
                                                             Date:   17 Sep 39
                                                             Page:    2  of 20
                 4.2.7.1.3  Where
                 The audit form should  have  a  space  to show  the location of the
sensor on a  tower,  including height.   A  sketch  is useful  to show the  relative
positions of  the sensing elements with respect  to possible  biasing influences,
such as the  tower,  other sensors and buildings.

                 4.2.7.1.4 Vhen

                 The date or dates when the  audit affected the system  operation
should be listed.   The time when the system or  a particular sensor was taken
"off-line" and  put  back "on-line" should be listed.  The  time, or time period,
when each datum value was taken is vital for  the comparison with the  system
output.  Implicit in this is the need  for the time  the auditor uses to be
correlated to the time the operator or the  system uses.   The auditor  should
rely on the  National Bureau of Standards station WWV for  correct time.  Battery
operated receivers, such as the Radio  Shack Time Cube, are  generally  available.

                 4.2.7.1.5  How

                 The audit form should  either  contain a copy of the method used
or reference the method number.  The audit  report should  contain copies of the
audit method used.   The methods should be detailed  enough to identify each step
in the acquisition of the audit value  and in  the conversion of the value to
units compatible with the system output.

       4.2.7.2  Wind Speed

       There are two general philosophies in  use by those who operate
anemometers  in  meteorological monitoring systems and networks.  The most coaaon
treats the system as a unit where the  sensor  and signal conditioner and
recorder are calibrated together.  The other, often employed by operators with
large numbers of anemometers, considers  the sensor  as a standard
interchangeable part.  In this case two  audits  are  necessary.  One to challenge
the sensor calibration method and the  other to  challenge  the system calibration
using a standard signal as a substitute  for the sensor.   A  full system audit
method from  sensor input to system output can be used as  a  challenge  for the
system operated with interchangeable sensors.

                 4.2.7.2.1 Sensor Control

                 The controlled condition is rate of rotation of the anemometer
shaft.  The  cup assembly or propeller  is removed for this challenge.  The audit
form should  provide space for fully defining  the transfer function used by the
operators  (usually supplied by the manufacturer).   This should include  the
relationship of rate of rotation (R, rps) to  wind speed  (U, m/s), rate of
rotation to  output volts (0, V) and rate of rotation to frequency (f. Hz), for
light chopper or a.c. generator types.  See 4.2.2.1.2 for a discussion of the
U=a+bR and U=bR types of transfer functions and how the constants "a" and "b"
are determined.  Some manufacturers provide the transfer  function in  the form
                              f = 26.439  (U  -  0.281)
which can  be converted to
                                U =  0.281  +  1.135R
once the number of pulses per revolution (30  in this case)  is known.

-------
                                                             Figure 4.2.7.1
                                                                   \
                                                                      I
                                                              Section  No.  4.2.7
                                                              Revision No.     G
                                                              Date:    17 Sep 89
                                                              Page:     3  of 20
                The rate of rotation can be  imposed
on the anemometer shaft in a number of ways.   If  the
method is to drive the shaft with a d.c. motor, the
number of revolutions of the shaft over a period  of
time is the data value.  That value divided  by  the
number of seconds in the time period gives average
rate of rotation R (rps).   The R is converted  to  U
by the transfer function and the U is compared  to
the system output in the same units for exactly the
same period of time.   If the system provides 5
minute average speeds, the count is for 5 minutes
with the start and stop times inclusive of the
system period.  If the d.c. motor is reasonably
constant (± 10%), a few seconds out of
synchronization over 300 seconds is acceptable.   The
period of time, however, must be exactly 300 seconds
which can be hand timed with a sweep second watch to
about ± 0.2 s.  If the system only reports hourly
averages, and cannot be changed to a shorter time,
samples of the signal conditioner output voltage  may
be used to estimate the system output.   Three rates
of rotation in addition to zero are recommended.
Since the important speeds are low speeds and not
full scale speeds, the use of simulated speeds  on
the order of 2, 5 and 10 m/s is acceptable.   Using
the transfer function above, these speeds are
simulated by R values of 1.51,  4.16 and 8.56 rps
(90.6,  249.6 and 513.6 rpm).  Figure 4.2.7.1 is an
experimental d.c.  motor drive used for this kind  of
audit.   Figure 4.2.7.2 is a second generation d.c.
motor drive capable of being powered by a D cell  and
a 9 volt transistor battery.
          If a propeller anemometer with a transfer
function of
                        U = 0.294R
is challenged, the speeds of 2,  5 and 10 m/s will be simulated by 6.9,  17,  and
34 rps (414,  1020,and 2040 rpm).   If the auditor  could generate five R  values,
1.5,  4,  7,  17 and 35 rps,  both cups and propellers could be challenged  at  three
meaningful  speeds plus zero.  See Figure 4.2.7.3  for a third d.c. motor system.
                                                                                 '.
                                                                              ya/t
                                                          Figure 4.2.7.2
Figure 4.2.7.3  An experimental 12-volt d.c.  motor and counter

-------
                                                              Section No.  4.2.7
                                                              Revision No.     3
                                                              Date:    17 Sep = 9
                                                              Page:     4  of 23
Climatronics cup
Teledyne-Geotech  cup
R.H. Young propeller
  Figure 4.2.7.4
     A simple d.c.  motor might be made to turr.
the shaft,  but the  key to the audit challenge is
the measurement of  the shaft revolutions.  A
light chopper and counter is a straight forward
approach to this measurement.  Hand switching
the counter for periods as short as 60 s will
produce better than one percent accuracy in
time.  If the light chopper produced 10 counts
per revolution, the count rate required for the
four R values mentioned above is 5 to 60 Hz, an
easy range for simple battery operated counters.
A system such as described above has the
advantage of independence from commercial power,
a significant advantage for some remote wind
systems.

     The controlled condition audit requires a
hands-on policy regarding the sensor.  The
measurements of the starting torque does not
require the sensor  to be connected to the
circuit.   It is possible, with proper equipment
and care, to install the speed challenge motor
on the sensor and operate that challenge with it
connected in its operating location.   Another
way to challenge an anemometer is to connect the
shaft to a synchronous motor.  The assumption
here is that the motor is running in
synchronization and the R value is therefore
known from the specifications of the motor or an
independent measurement of its rate of rotation.
Figure 4.2.7.4 shows three anemometers coupled
to a synchronous motor through a universal
coupler.

     The last measurement is the time constant.
With the simulated  speed on its highest rate,
and with a meter on the output voltage of the
signal conditioner, turn off the d.c.«otor and
measure the time it takes to reach the value of
the simulated speed minus 637. of the simulated
speed.  Examples of wind speed audit procedures
and forms are found in Figures 4.2.7.5 through
4.2.7.8.

-------
                                                        Section No   4.2.7
                                                        Revision No.    0
                                                        Date-   17  Sep  89
                                                        Page-    5   of  20
            Cup Anemometer - MSI method CAOO3  (version e'1

       This method provides for a comparison of the transfer  function
       used with the system to the output of the system.  This  is  done
       by causing the anemometer shaft to turn at a known rate  of
       rotation and observing the output.  The means of turni.ng the
       shaft and measuring the rate of rotation are provided  by the
       auditor and are completely independent of "t'fie operating  system.
       The method does not challenge the transfer function.   This  can  be
       done best with a wind tunnel test.

       The report form for this method includes space for an  optional
       determination of starting torque and system time constant.  The
       torque measurement may be used as an indication of bearing
       condition and hence starting threshold of the anemometer.   The
       time constant is of use if turbulence is measured.

       CAO03-A  Remove the cup assembly.  Mount a coupler to  the
       anemometer shaft.  A 1/8" shaft is required.  If the anemometer
       does not use that size or it is not accessible, an interface
       fitting will be required.  Clamp the drive motor to the  support
       column of the shaft so that the coupler is engaged with  the drive
       wheel.  Determine if the cup assembly turns the shaft  in a  clock-
       wise or counter clockwise direction, when viewed -from  above.
       Clockwise is common and is used on the form.  Operate  the drive
       motor at two speeds (find the desired rps from the transfer
       •function) which are important to the application of the  wind
       speed data.  Use a time period synchronous with the system
       output.  An average of one minute or longer is required.  If  the
       system provides only instantaneous samples of output volts, tale
       12 samples over a two minute period and use the average  of  the
       samples to compare with the average rate of rotation measured.

       CAO03-B  This method requires that the system be operating  with
       all cables in place (short jumper cables may be used with CAOO3-A
       to allow simultaneous access to the anemometer end the signal
       conditioner for those systems where these two parts are  at  some
       distance away).  At least a rero rate of rotation must be
       measured (or observed) with the anemometer in place, the cup
       assembly removed and the shaft taped to assure non—rotation.  A
       second observation mav be either a motor driven measured rate of
       rotation for the operating period of the system or a natural  (un-
       measured) non-zero operation to assure that signal reaches  the
       signal conditioner when the system is in operating position.  The
       assumption with the later choice is that if the signal is
       transmitted at all it will be properly simulated in method  A.
       This is more likely true with pulse trains than with generator
       voltages.
Figure 4.2.7.5  Audit  method for a cup anemometer

-------
                                                         Section No. 4 2.7
                                                         Revision No.    0
                                                         Date:   17 Sec =?
                                                         Page:    6  of 20
                 PERFORMANCE AUDIT REPORT by  	      CAO03

      MEASUREMENT SYSTEM - Cup anemometer
            System number ______________________-.—	—•	
            Sensor        __________________————.	
            Cup  assembly
            Location
            Signal  conditioner 	
            Data Channel	
      DATE 	/	/	TIME of-f  line	on line	test start

      TRANSFER FUNCTION:   
-------
                                                     Section No. 4.2.7
                                                     Revision No.    0
                                                     Date:   17 Sep 89
                                                     Page:    7  of 20


     Fixed A>; i s Propeller - MSI method  FAF001  (version 6/l/B-»

This method provides •for a comparison of  the transfer -function
used with the system to the output of the system.  A separate
•form is provided for W (vertical component) since a different
transfer function is often used for this  direction than is used
for U and V.  The method causes the propeller shaft to turn at  a
known rate of rotation while observing  the output.  The means of
turning the shaft and measuring the rate  of rotation are provided
by the auditor and are completely independent of the operating
system.  The method does not challenge  the transfer function.
This can be done best with a wind tunnel  test.  The sign
convention used with respect to clockwise and counter clockwise
is that of the system being challenged.   Differences are always
calculated by subtracting the audit challenge value from the
system output.  Arithmetic convention is  followed even though the
minus sign is used as an indicator o-f direction.  For example.
the difference between a -1.5 mps audit challenge and a -1.3 tops
system output is +0.2 mps even though the system underestimated
the speed  (a negative error) with respect to the audit value in
the "-" direction.

The report form for this method includes  space for an optional
determination o-f starting torque and system time constant.  The
torque measurement may be used as an indication of bearing
condition and hence starting threshold of the propeller.  The
time constant is of use if turbulence is  measured.

FAPO01-A  Remove the propeller.  Mount a  coupler to the propeller
shaft.  A  1/8" shaft is required.  If the propeller does not use
that size or it is not accessable. an interface fitting will be
required.  Clamp the drive motor to the support column o-f the
shaft so that the coupler is engaged with the drive wheel.
Operate the drive motor in both a clockwise and counter clockwise
direction, when viewed from in front of the propeller.  Operate
the drive motor at two speeds (find the desired rps from the
transfer function) which are important to the application of the
wind speed data.  Use a time period synchronous with the system
output.  An average of one minute or longer is required.  If the
system provides only instantaneous samples of output volts,  take
12 samples over a two minute period and use the average of the
samples to compare with the average rate  of rotation measured.
If a synchronous motor is used as the drive motor, an instan-
taneous sample voltage may be used.  Some evidence of the synch-
ronous operation of the motor is required.

CAO03-B  This method requires that the system be operating with
all cables in place (short jumper cables may be used with FAPOO1-A
to allow simultaneous access to the propeller and the signal
conditioner -for those systems where these two parts are at some
distance away).  At least a zero rate of  rotation must be
measured  (or observed) with the propeller shaft in place,  the
propeller  removed and the shaft taped to  assure non-rotation.   A
second observation may be either a motor driven measured rate of
rotation for the operating period of the  system or a natural (un-
measured)  non-zero operation to assure that signal reaches the
signal conditioner when the system is in operating position.

Figure 4.2.7.7 Audit method for a propeller anemometer

-------
                                                       Section No. 4.2.7
                                                       Revision No.    0
                                                       Date:   17 Sep 89
                                                       Page:    8  of 20
              PERFORMANCE AUDIT REPORT by  	    FAP001W

   MEASUREMENT SYSTEM - Fixed axis  propeller
        System number 	
        Sensor
        Pr-opeller
        Location
        Signal conditioner
        Data channel    	
   DATE  	/	/	 TIME o-ff line	on  line	test start	

   TRANSFER FUNCTION: 1 rps = 0.294 mps (3 pulses per revolution)
                         CW-2.53*4 = m/s

   TEST  RESULTS

     FAPOO1W-A      - challenge  speed -    output   dif-ference

                   time   revs,   rps    mps    mps    mps     /.

         O       	                    0	

       "d.c."

         S CW

         S CCW

         F CM

         F CCW

       "sync"      time    rps     mps   volt    mps    mps      '/.

         S CW    	   5.0OO   1.47	

         S CCW   	   5.000  -1.47 	

         F CW    	   30.OO   8.82	

         F CCW   	30.OO  -8.82	

         Torque:	Oz.-In.  cw,	Oz.-In. ccw. T. Const.	s.

      FAPOO1W-B            expected    observed      difference
                   time        mps         mps        mps     *>
         O

       test
Figure 4.2.7.8  Audit form for the propeller anemometer method

-------
                                                             Section No. 4.2.7
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    9  of 20
                4.2.7,2.2  CIS Method
                The collocated transfer standard (CIS) method for wind speed
involves mounting a carefully calibrated anemometer in the vicinity of the
subject anemometer being audited.  The CTS should have certificates tracing its
calibration to NBS or some other standard facility.  If the ASTM (1984) method
for comparability is being used, the CTS needs to be within 10 m of the subject
anemometer in the horizontal and the lesser of 1 m or H/10, where H is the
height above ground in meters, in the vertical.  It is important to site  the
CTS to be representative of the flow at the subject anemometer.  Mutual
interference should be minimized through siting and through editing out data
where the direction shows the wind passing through one to reach the other.  The
accuracy potential of the CTS method is based on data taken in 1982 at the
Boulder Atmospheric Observatory  (BAO) and published by Finkelstein et al.
(1986) and Lockhart (1988).  The anemometers for this study were spaced about 5
m apart.   The closer together they are in the horizontal the larger the
direction sector of mutual interference.

                The best situation for CTS auditing is one in which both
anemometers are connected to the auditor's data logger.  The element of the CTS
audit Is the difference in speed calculated by subtracting the CTS speed  from
the subject speed.  The method requires a sufficient number of simultaneous and
independent differences.  A simultaneous difference is one where the time
between sampling each anemometer output is less than O.lT, where T (s) is found
by dividing the distance constant, D (m), by the wind speed U  (m/s).
Independence is achieved when the time between sampled pairs is larger than 4z.
For example, assume the subject anemometer has a distance constant of 5 m and
the CTS has a distance constant of 1 m.  If the wind speed is about 3 m/s. *
will be 5/3 = 1.7 s for the subject and 1/3 = 0.3 s for the CTS.  Simultaneous
samples will exist when the sampling rate of the data logger is less than the
shortest O.lT or 0.03 s in this example.  Most data loggers are fast enough for
the example.

                Independence is achieved when the time between successive
sample pairs is long enough.  In the example above, 4r is 20/3 = 6.7 s for the
subject anemometer and 4/3 = 1.3 s for the CTS.  If the CTS logger is set for
one sample every 10 s, the data will be independent at 3 m/s.   The ASTM method
defines minimum sample size in terms of the resolution of the measured or
reported speed (assume 0.1 m/s for the example, a recommended resolution) and
the standard deviation of the series of differences.   It takes 900 times  the
variance of the series to provide the minimum number.   If the two sensors are
well sited and properly operating, the variance will be small.   The BAO data
showed the standard deviation of the difference to be less than 0.2 m/s
(variance of 0.04).   This should be the minimum condition for a good CTS data
set.  If the variance is 0.04. the minimum sample size is 900 x 0.04 = 36.

                Assume the subject anemometer produces a scalar average speed
every 15 minutes and it is not possible to wire the output into the auditor's
data logger.  The CTS uses 90 samples,  one each 10 s,  to assemble its
concurrent 15 minute scalar average.  At this point,  one can take one of  two
paths.  One Is to assume that the subject is operating well,  has a short enough
distance constant and is likely to agree well with the CTS on a sample to
saaple basis;  not perfectly because a large bias can still have a small
variance.   There is no way to verify this assumption unless the audit results

-------
                                                             Section No. 4.2.7
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:   10  of 20


show good agreement.  Based on this  assumption, each  15 minute value is a data
point with a sufficient  sample size  to compare  the anemometers.   Each datum
point can be used  in a linear regression analysis to  define the subject
anemometer's accuracy as a function  of wind speed, if enough dynamic range
exists in the data period.   In just  a few hours, then, the CIS method with this
assumption will produce  a measure of accuracy.

                The other way is to  treat each  15-minute average  as a single
sample of 15-minute averaged data.   A 9 hour period will provide  the minimum
sample size of 36.  This is an awkward period of time.   It takes  some time to
install the CTS and many auditors work 8 hour days like other people.   There  is
no such thing as  too many samples.   Added time  usually enlarges  the dynamic
range for the audit.  The optimum CTS audit goes for  something  like 24  hours.
one diurnal cycle.  For  an example of this CTS  audit, look at the data  listed
in Table 4.2.7.1.
                Each value is a 20-minute scalar average.  The  subject
anemometer  is a Climatronics F460 cup (C-V-W).  The CTS  is a Young Propeller
Vane  (P-V-U)  located 5 m away.  Both are at a height  of  10 m.   The standard
deviation of  the  CTS sensor is shown for each 20-minute  period.   The difference
between the subject anemometer and the CTS is shown  in  the column headed by
"Y-X."  Notice  that  the  average difference is a small -0.12 m/s and  the
standard deviation of  the difference series is  0.10,  half of the maximum
criteria.   To express  the accuracy of the subject  anemometer with respect to
the CTS, a  linear regression was run.  The constant  of  0.02 m/s says  there is
no bias of  significance.  The X Coefficient of  0.96  says that  there  is  a 4 '/.
underestimate of  speed at all speeds.  The best fit  straight  line through the Y
points is calculated by multiplying the X  (true)  value  by  the coefficient and
adding the  constant.   The residual error is then  found  by  subtracting  the X
value from  the  estimated Y value, Y* , (column headed by Y'-X).   Notice  that  the
average difference is  0.000, as it must be, and the  standard deviation  is
lower, 0.07 m/s.   The  audit report for this subject  anemometer  would  report  the
error of 4  '/.  in the  slope of the subject transfer function.   (NOTICE: This is
an example.   The  CTS is arbitrarily selected.  It is possible  that in  this case
the CTS was 4 '/. high or they were each off 2 '/..  That is not  important  for this
example and the speeds are so close that this analysis  was  not  included in the
BAO experiment.)   These data are shown graphically in Figure  4.2.7.9 and Figure
4.2.7.10.   The  anemometer data are shown on an XY plot  with the best  fit
straight line through the 72 points.  Figure 4.2.7.10 shows an XY plot  of the
residuals.   Also shown  is  the normalized turbulence,  
-------
                                           Section No. 4.2.7
                                           Revision No.    0
                                           Date:   17 Sep 89
                                           Page:   11  of 20
Table 4.2.7.1  Simulated Audit Data (BAD 1982)
r/32
Sate/
TlK

4 iCCO
4 1520
4 1040
4 HOC
4 !!20
1140
1200
!22C
1240
1300
4 ISO
4 1340
4 1400
4 14:0
4 1440
4 1500
4 ISO
4 1540
4 1600
4 !e20
4 1640
4 1700
* 1720
4 1740
4 1300
4 1E20
4 :S40
* !*>0
< :?20
l <:(0
* :-xw
4 2520
4 2^0
4 2190
4 2120
4 21*0
* zz-oo
4 1220
i r*o
* 2:-w
* ~:0
< "*0
5 0
5 :o
5 *0
5 :co
¥
Subject
Speed
(l/S)
3.701
2.972
3.5S7
2.660
2.463
2.376
3.210
5.511
4.546
4.261
4.157
3.127
2.364
2.451
4.176
8.281
7.673
6.371
5.569
3.346
8.338
7.236
5.425
3.989
4.399
4.407
3.847
3.941
4.344
4.300
2.907
2.861
2.049
1.747
3.796
4.065
3.635
4.221
4.097
4.037
3.952
3.559
3.449
3.179
3.340
4.393
I
CIS
Speed
(a/s)
3.729
2.979
3.555
2.605
2.365
2.399
3.344
5.623
4.722
4.555
4.455
3.250
2.433
2.498
4.270
8.601
7.893
6.600
5.976
3.413
8.693
7.486
5.538
.111
.617
.615
.012
.121
4.474
4.474
3.035
3.156
2.221
2.015
3.957
4.198
4.031
4.376
4.261
4.194
4.135
3.717
3.530
3.290
4.040
4.617
s
CTS
Sigaa
(I/S)
1.047
2.441
2.780
0.844
1.412
0.883
1.951
3.230
1.115
1.319
1.348
1.120
1.557
1.039
5.887
2.564
1.635
1.688
2.8*7
3.537
4.531
1.503
0.751
0.233
0.520
0.930
0.345
0.291
0.151
0.363
0.442
0.154
0.022
0.190
0.246
0.215
0.137
0.206
0.232
0.194
O.I4«
0. 109
0.123
0.035
0.199
0.113
s/I
(•/s)
o.:a
0.82
0.73
0.32
0.60
0.37
0.55
0.57
0.24
0.29
0.30
0.34
0.64
0.42
1.38
0.30
0.21
0.26
0.48
1.04
0.52
0.20
0.13
0.06
0.18
0.20
0.09
0.07
0.03
o.oe
0.15
O.C5
0.01
0.09
0.06
0.05
0.03
0.05
0.07
0.0:
0.03
O.C3
0.03
0.01
O.C5
0.02
Y-I
(B/S)
-0.03
-0.01
0.00
O.C5
0.10
-0.02
-0.13
-0.11
-0.18
-0.29
-0.30
-0.12
-0.07
-0.05
-0.09
-0.32
-0.21
-0.23
-0.41
-0.07
-0.36
-0.25
-0.16
-0.12
-0.22
-0.21
-0.16
-0.18
-0.13
-0.17
-0.13
-0.29
-0.17
-0.27
-0.16
-0.13
-0.15
-0.16
-0.16
-0.16
-0.18
-0.16
-0.13
-0.11
-0.20
-0.22
r-i
(•/s)
-0.11
-0.10
-0.13
-0.14
-0.17
-0.06
0.02
-0.10
-0.00
0.12
0.13
0.01
-0.01
-0.04
-0.06
-0.02
-0.10
-0.03
0.18
-0.05
0.01
-0.04
-0.05
-0.03
0.05
0.04
0.02
0.03
-0.04
0.01
0.02
0.15
0.10
0.21
0.02
-O.C2
-0.00
-0.01
C.01
0.00
0.03
G.02
0.00
-y.Ol
0.05
0.05


5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
C
ij





120 3.822 3.955 0.881
140 2.198 2.219 0.520
200 1.630 1.742 0.059
220 1.484 1.510 0.576
240 1.599 1.624 0.377
300 2.569 2.697 0.076
320 1.662 1.696 0.341
340 1.769 1.934 0.190
400 1.054 1.114 0.459
420 1.471 1.455 1.217
440 2.131 2.236 0.438
500 1.338 1.419 0.680
520 3.315 3.503 0.180
540 3.060 3.212 0.220
600 1.992 2.074 0.379
620 1.765 1.836 0.336
640 0.920 0.396 0.072
700 0.636 0.523 0.280
720 1.809 1.856 0.422
740 1.571 1.577 0.173
800 2.058 2.115 0.292
820 1.910 2.000 0.519
340 1.181 1.165 0.345
900 2.880 3.013 0.891
920 4.236 4.402 0.537
940 3.339 4.000 1.752
Average
Standard deviation



0.22
0.23
0.03
0.38
0.23
0.03
0.20
0.10
0.41
0.34
0.19
0.48
0.05
0.07
0.18
0.18
0.08
0.54
0.23
0.11
0.14
0.26
0.30
0.30
0.12
0.44





-0.13
-0.02
-0.06
-0.03
-0.02
-0.13
-0.03
-0.17
-0.06
0.02
-0.10
-0.08
-0.19
-0.15
-O.OB
-0.07
0.02
0.11
-0.05
-0.01
-0.06
-0.09
0.02
-0.13
-0.12
-0.16
-0.125
0.100



-0.01
-0.05
0.01
-0.01
-0.02
0.04
-0.01
0.11
0.04
-0.05
0.03
0.04
0.06
0.04
0.02
0.02
-0.04
-0.11
-0.01
-0.04
-0.01
0.03
-0.04
0.03
-0.05
0.01
0.000
0.069

Regression Output:
















Constant
Std Err of Y Est
R Squared
No. of Observations
Degrees of Freedos

\ Coeff. 0.9576
Std Err 0.0048








































0.0233
0.0699
0.9982
72
70












-------
     SIMULATED  CTS  AUDIT             \l
Propeller Vane (P-V-W) as Collocated Transfer Standard
                                                      Section No. 4.2.7
                                                      Revision No.    0

                                                                   20
         Regression Analysis
         Y = 0.023 +  0.96 X
                                           Data: Boulder Experiment
                                                9/4-5/1982
                                           72 - 20 minute averages
                    CTS  (P-V-W) Wind Speed (m/s)
Figure 4



In
c
xl>

O 0-2 -
o
(T. 0 -
.2.7.9 XY plot of simulated wind speed audit data
KEY: * = Sigmo/Speed
A
A .A 	 	 - 	
^ A A A
A A
. * . A , A 	 _
^..y AA & ^ A


T +. n * KEY: + = C-V-W residual
"^ ^ ,
* + + ^V* + ++" ^ ' ++ * + + ^
•o
1 B S
Q.
l^^
CO
00^
O.O Q>
CO
°-6oo
«*—
On o
" r£



5:     0           2          4          6          8

                   Wind Speed  (m/s) of CTS  (P-V-W)

Figure 4.2.7.10  /tesiduaJ analysis o/ a sjjnulat*?d speed audit
                                                  10

-------
                                                             Section No. 4.2.7
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:   13  of 20
                4.2.7.2.3  "V Anemometers
                Some stations measure the vertical component of the wind, with
an anemometer sensitive only to the vertical component of the wind.  A
vertically mounted helicoid propeller, or "V" propeller,  is the most common
instrument for this measurement.  The same audit methods can be used as are
used on a propeller anemometer.  A synchronous or d.c. motor will challenge the
rate of rotation vs. wind speed and a torque device can be used to find the
starting threshold.  The common manufacturer's recommended practice is to use a
different transfer function for the W propeller than the one applied when the
same propeller is used for vane-oriented speed or for the N-S and E-W
components of a UVW anemometer.

            4.2.7.3 Wind Direction

                4.2.7.3.1 Sensor control

                The first thing to do on a performance audit of a direction
vane is to record the as-found orientation value.  Have the operator hold the
vane so that it points to or from (whichever is most accurate for aiming) the
distant orientation target.  Verify the alignment by viewing the vane and
target from the ground.  Move back away from the tower or mast on a plane which
passes through the sensor and the target and verify that the vane is in the
plane also.  Field glasses or a theodolite can help make this sometimes
difficult observation.  The vane must be held steady or clamped until a
constant output exists for a few minutes.  Record this value.

                The controlled condition for a wind vane is a relative position
of the vane with respect to the sensor housing.  There are several ways to
lapose a series of known relative positions on the vane-sensor combination.
They vary in effective accuracy.  It Is critical to know the time constant of
the direction circuit BEFORE starting the performance audit.  It can be
measured by setting the vane to a known direction, simulate a wind from 090°
holding the vane steady until the 090  (or voltage equivalent) output is
steady.  Move the vane quickly  (< 1 s) to 270  and measure the time constant of
the system.  Assume that a time constant of 3 s is measured.  Table 4.2.7.2
shows the change in output angle and voltage (assuming a 540° format and SV
output) as a function of time.

                     Table 4.2.7.2 - Time Constant Effects
Time Time
Constant Angle
(No.) (sec.)
0
0.2
1
2.3
3
4.6
6.9

0
0.5
3
6.9
9
13.8
20.7

Vane 	 Output 	
Angle Angle Error (540@5)
(deg.) (deg.) (deg. ) (volts)
090
270
270
270
270
270
270

090
106
204
252
261
268
270
(after
0
164
66
18
9
2
0
Fritschen
0.333
0.981
1.889
2.333
2.417
2.483
2.498
and Gay,
Change
( '/. )
0.0
9
63.2
90.0
95.0
99.0
99.9
1979)

-------
                                                             Section No. 4.2.7
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:   14  of 20
                Notice that in this example a 180  shift requires waiting 20
seconds for the reading to be representative of the new position.  If a 90
shift is used, 14 seconds will provide an output within 1  of the final value.
If measurable time constants are found, suggest to the operator  that the
manufacturer be called about steps which might be taken to modify the circuit  to
a •inimuM time suitable for 60 Hz noise filtering.

                The least accurate method for challenging the relative position
accuracy of a wind vane is to point the vane in various directions while still
nounted on the tower.   This can provide positions related to external objects
rather than constant angle changes.  It is estimated that the accuracy of this
Method is two to five degrees, with the exception of a parallel  alignment.  The
tail vane can be located parallel to a cross arm to within one degree, and held
parallel on a calm day.

                A second method puts the operating sensor in a controlled
situation like a room where the electronics are located.  The sensor can be
placed at the center of a template with radial lines every 60
re oriented to the template and the vane
noved and clamped when the vane is parallel
-.o the radial line.   If care is taken to
avoid parallax errors (non-parallel or
non-perpendicular observations) this method
can provide relative accuracy on the order
:f one degree.
                The best method replaces
:ae vane with a fixture with the capability
:f holding the shaft in fixed positions
-•ith respect to the sensor housing.
Fixtures of this type can provide
repeatable position accuracy of < 0.1 .
Figure 4.2.7.11 shows such a device.  A
different application of this precise
nethod uses a theodolite base as the mount
::r the sensor.  With the vane or vane
substitute held in one position, the base
:=n be rotated in very accurate steps.
"r.eodolite worm gear assemblies divide a
rircle in whole degrees with a vernier
sijustment with 0.1 degree index marks far
r-^ugh apart to allow easy interpolation to
I  32 degrees,  a resolution wasted on the
application of wind direction measurement.

                The audit report form
=-ould contain the transfer function used
:: convert output voltage to azimuth
;=grees.   This may include a 540 format
--.ere azimuth values greater than 360 are
re-iuced by subtracting 360.   The report
:~TT:H should also contain the challenge
pr-gression used by the selected method.
         The sensor  can
 Figure 4.2.7.11

Figure 4.2.7.12

-------
                                                             Section No. 4.2.7
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:   15  of 20


For example. Fig. 4.2.7.14 and Figure 4.2.7.15.show an audit methodoand audit
form for wind direction which specifies 16 relative angles,  each 60  from the
last one,  in a clockwise rotation for 420° followed by a counterclockwise
rotation of 480°.  This tests a 540 format,  provides four samples at 180°
(duplicates from each direction) and a duplicate counterclockwise 240° pair.
The report of this series describes the range of relative error resulting from
the shaft position measurement of the sensor (see 4.2.2.2.2.3 for an example).

                The starting threshold of the bearing and transducer assembly
should be Measured by some method (see 4.2.2.2.1.2 and 4.2.2.2.1.3).  If the k
value is not available to convert the torque to threshold speed at some
accuracy angle, the operator should be requested to ask the manufacturer to
provide it for the next audit.

                The bearing to the orientation target should be independently
challenged with a method capable of better than compass accuracy.   A theodolite
is ideal for finding the bearing to other distant objects.   A solar observation
is recommended (see 4.2.4.3.2).

                The last activity of the sensor control audit is to repeat the
orientation test described above for the as-found value.  The as-left value
will represent any changes the operator may have made and the new orientation,
if the sensor was not keyed for orientation.

                4.2.7.3.2 CTS Hethod

                There is no technical need for a CTS audit of direction.  No
new information is added by this method to that gained in the sensor control
method.   As a parallel example to the simulated CTS speed audit, data from the
same period of time from BAO is shown here,  structured as a simulated CTS
direction audit.   Table 4.2.7.3, sorted for ascending CTS direction, shows the
20-Binute average direction for both the CTS (Young Propeller Vane, P-V-W) and
the subject (Climatronics F460, C-V-W).  Also shown is the 
-------
                                                   Section No.  4.2.7
                                                   Revision No.     0
                                                   Date:    17 Sep 89
                                                   Page:    16   of 20
Table  4.2.7.3  Simulated CTS Direction Audit Data
"82
Date/
Ti«e

5
5
5
S
4
4
4
S
4
4
4
4
S
4
4
4
*
4
5
5
4
4
S
S
5
4
5
S
S
4
4
4
S
S
3
S
S
4
4
5
5
S
3
5
5
4
a
5
j
*
3
x
5
S
4
A
X
m
A
A
i
V
K
V
4
X
i
5
f.
e
X
X




740
720
114O
44O
:ooo
1B4O
1B2O
1O4O
22OO
2120
2J4O
19OO
11OO
1620
1720
170O
1740
16OO
1O2O
62O
164O
192O
SCO
20
O
2320
112O
1OO
40
23OO
2220
234O
9OO
120
92O
6OO
420
2240
194O
520
22O
540
820
300
32O
210O
1020
940
114O
2040
200
202O
4OO
34O
130O
2OOO
1320
140
154O
1600
1200
840
1200
BOO
124O
122O
1540
240
1S2O
1OOO
14A<.
1500
V
Subject
Direction
Cdeg. >
11O. 72
126.18
136.01
149.21
170.93
171.89
173.12
176.16
175.63
179.81
1BO.49
181. S5
183.77
184.24
186.51
186.90
187.74
189.05
189.53
191.87
193. Ol
194. Ol
196.86
197.68
199.46
2O1.43
2O2. 16
2O3.33
205.00
2O5.77
205. 97
2O7.6S
2O7. 16
2O9.2S
212.67
214. O9
214.12
216.04
217.59
223.97
229.63
227.38
23O.43
232.28
239.91
237.55
239. 85
239.28
1-42.97
247.28
253. 86
257. 5O
260.27
264.41
2o4.B6
267.01
269.87
273.76
274.96
275.77
277.95
289.50
2<»9. 15
300. 55
3O1.51
31O.99
312.43
318.79
T24.82
7-49.62
359. 13
3e0.42
X
CTS
Direction
(0*9.)
108.58
121. 6O
134.38
148.15
170.74
171.50
173.71
174.52
179.24
17B.59
179.64
180.68
182.95
183.74
185.29
185.82
187. O7
187.82
19Q. 19
191. Ol
191.43
192.68
193.O1
197.17
198.88
2O1.11
2O1.37
2O2.99
2O4.5O
205.64
206.42
2O7.02
207.29
2O9.08
212.72
214.27
215.12
215.95
217.16
224. 12
225. 53
228.04
229.87
232.98
237.14
237.75
240. O6
241.26
244.21
248.22
254.16
257.72
257. 8O
263.64
264.68
266. 10
27O.39
273.13
274.04
275.35
277.67
289.42
297.87
298. Ol
3OO. 18
308.24
31O.OO
315.62
322. 49
345.54
T55.74
356. 8C>
•
CTS
SlQM
(deg.)
39.34
20.76
87.87
6.50
27.02
10.17
3O.55
16.28
7.62
4.04
10.87
9.OO
15.26
6O.97
6.IO
8.98
6.66
17.58
27. 04
32.37
9.24
9.10
30.40
4.44
5.99
6.94
22.84
3.45
5.35
6.21
7.9O
5.88
11.83
9.57
9.31
B.OO
20.18
5.86
B.76
6.29
17.04
9.38
28.91
8.80
25.28
35.85
42.30
o3.08
29.30
12.10
75.26
11.45
20.16
12.57
23.11
29. 16
22.09
43.68
34.O"
21.27
27.46
29.47
11.74
65.04
22.86
17.77
10.46
21.03
12.26
o".6o
30.58
20.56
Y-X


(deg. >
2.13
4.57
1.63
1.O5
0.18
O.39
1.42
1.63
0.39
1.22
O.8S
O.87
0.82
0.50
1.23
1.08
O.67
1.22
-O.66
0.85
1.59
1.32
3.85
0.52
O.58
O.53
O.79
O.34
0.50
0.13
-O.45
0.63
-0.13
0.16
-0.05
-O. 17
-l.OO
O.08
0.42
-0. 15
4.1O
-O.66
0.56
-O.70
2.77
-O. 19
-O.20
-1.98
-1.25
-O.94
-0.30
-O.22
2.47
0.77
O.19
0.91
-O.51
0.63
O.92
O.42
0.28
O.OB
1.28
2.55
1.35
2.75
2.43
3.17
2.53
4.08
3.4O
3.62
                                        Average                       O.91
                                        Standard deviation             1.34

                                                 Regression Outputi
                                        Constant                   -O.22090
                                        Std Err of  V Est             1.35045
                                        R Squared                   0.99939
                                        No. oi Observation*              72
                                        Degrees of  Freedom               70
                                        X Coefficient(s)
                                        Std Err of Coef.
1.OO498
O.OO297

-------
                      SIMULATED  CTS AUDIT
              Propeller Vane (P-V-W) as Collocated Transfer Standard
                                                           Section No. 4.2.7
                                                           Revision No.    0
                                                           Date:   17 Sep 89
                                                           Page:   17  of 20
   Q>
  ^ 360 -
   c
   o
  "o
   

                The sensor control method is used for bivane auditing.   The
vertical  part of the  bivane operation is treated in  the same way as the
horizontal  part, except  different fixtures are used.   Special fixtures  are
required  for each bivane design, but the principle is the same.   A relative
zero point  is set when the fixture is attached to the bivane.   That point is
where the vane shaft  is  perpendicular to the vertical axis of the sensor.  Fron
this starting point where  the output should be the equivalent of 0 ,  the vane
is held in  15° steps  until its physical limits are reached,  both tail up ( + )
and tail  down (-).  Threshold is very hard to measure on a bivane because of
the static  balance conditions of the vane.  If the vane is perfectly balanced
and its remains where ever it is physically moved, a force gage measurement at
sone distance from its axis of rotation will yield the starting threshold Just
as the vane begins to move.

       4.2.7.4 TURBULENCE  OR 
-------
                                                             Section No.  4.2.7
                                                             Revision No.     0
                                                             Date:    17 Sep 89
                                                             Page:    18  of 20


       The challenge should  be  realistic or at least within some realistic
range.   The challenge must take into consideration the wave shape of the
variable direction imposed on the system in calculating the true sigma value
with which the output will be compared.  The effective time constant of the
direction system, calculated from the delay distance of the sensor and some
nominal wind  speed important to air pollution applications, should define the
maximum frequency used  in the sigma challenge.

-------
                                                         Section No. 4.2.7
                                                         Revision No.     0
                                                         Date:   17 Sep 89
                                                         Page:   19  of 20
        Wind Vane - MSI method WVOO4 (version 8/1/84)

   This method describes the relative performance of  the wind vane
   as a shaft-position transducer  and the orientation of the
   transducer with respect to true North.   The former is.done with a
   fixture, part of which is mounted to the transducer body and part
   mounted to the shaft in place of the vane.   The latter" requires a
   determination of true North (see MSI method SN008)  and a setting
   of the transducer relative to that orientation.

   The report form for this method includes space for the optional
   method to define the "open space" where relevant to the sensor
   and application.  Also there is space to record starting torque
   measurements and system time constant estimates for turbulence
   measurement.

   WV004-A  Remove the wind vane assembly (vane,  shaft and countei—
   weight).  A 1/8" shaft is required.   If the sensor does not use
   that size or it is not accessible,  an interface fitting is
   required.  Mount the disc on the vertical shaft.  Mount the clamp
   to the support column for the shaft  so that the pin engages the
   disc and the disc is free to move when the pin is  withdrawn.  Set
   the fixture parts with the pin  in the 180 degree hole.  Rotate
   the clamp until the output indicates 180, either by equivalent
   voltage or digital  printout.  Since  this is a  position
   measurement, the challenge is constant and instantaneous values
   may be used, being sure to react to  the needs  of the time
   constant for stable readings.   Move  the disc (vane substitute) to
   the following positions taking  data  at each point:   12O, 060,
   360,  3OO, 240,  180,  120.  190, 24O,  3OO,  36O, 060,  12O, 180. and
   240 degrees.  This moves the "vane"  42O degrees counter clockwise
   and then 480 degrees clockwise  to test "540" strategies for the
   angle discontinuity.

   WV004-B  Define the "open space" for 360 degree potentiometer
   transducers.  Install  the index  line fixture to an appropriate
   position with respect to the protractor mounted to the disc.
   Disengage the pin.   Rotate the  disc  until the  output changes from
   maximum voltage to five degrees less than maximum.   Record the
   angle to 1/2 degree resolution.   Rotate the disc back toward
   maximum voltage and record the  angle when maximum  is first
   reached.   Rotate the disc until  the  output  changes from minimum
   voltage to five degrees greater  than the minimum.   Record the
   angle.   Rotate  the  disc back toward  the minimum voltage and
   record the angle when minimum is first reached.

   WV004-C  After  having found  a distant target of known .direction
   (see  SNO08),  set the vane so that the direction is the output  of
   the sensor.   Clamp  the vane  to  the shaft support tube"so that  the
   output stays constantly correct,  even with  light wind forces on the
   vane.   In heavy wind,  a fixture  replacing the  vane is required.
   Place the sensor in its mount and rotate the sensor body until
   the vane counterweight points to the target.  Clamp the sensor in
   place,  check the output and  remove the vane clamp.   Record one
   system data  point  with all cables in place  and the sensor clamped.


Figure 4.2.7.14  A method  for auditing a wind direction sensor

-------
                                                       Section No. 4.2.7
                                                       Revision No.    0
                                                       Date:   17 Sep 89
                                                       Page:   20 of 20
             PERFORMANCE AUDIT REPORT bv             _        WV004
  MEASUREMENT SYSTEM - Wind vane
       System number   .__
       Sensor	
       Vane          	
       Loc at i on	
       Signal  conditioner
       Data  channel  '	
  DATE 	/	/	TIME off line	on line	test start	

  TRANSFER FUNCTION (volts per degree) 	
                     discontinuity strategy	

  TEST RESULTS

    WVOO4A      (dif. » deg. - set)

      set  volt deg. dif.  set  volt deg. dif.  set  volt deg. dif.

      ISO	24O	300  	

      120	180  	360  	.	

      06O	120	06O

      360	180	120  	

      3OO  	  24O  	  180  	

                                                 24O  	

    WVOO4B          1                         2

      volt deg. angle-dif.      volt deg. angle dif..  abs(i1-2)-360)
      Toroue:	 Oz.-In. cw,	Oz.-In. ccw, T. Const.	s.

    WVOO4C           1                            2

      time     expected    obs. before    obs. after      difference

             volt  deg.   volt   deg.    volt   deg.     (2-1) deg.
Figure 4.2.7.15  A form for the wind direction audit  method

-------
                                                             Section No. 4.2.8
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    1  of  1

4.2.8  ESTIMATING ACCURACY AND PRECISION

       4.2 .• 8.1 Measurements

       Section No. 4.1.5 contains a detailed discussion of methods of
estimating accuracy, precision and bias using wind speed and wind direction as
examples.   That material will not be repeated here.  The measurement process
begins with an instrument which has some element or part which is sensitive to
the variable of interest.  If interest is in air flow with respect to the
surface,  there are two variables, wind speed and wind direction.  Each part of
this section discusses various aspects of how the measurements might be made.
calibrated, operated, maintained and documented to support a claim of
measurement validity.  It is recognized that Instruments are usually parts of
data systems with sampling, processing and summarizing routines designed to
produce final elements of a data base to be used for some application.   The
earlier parts of this section were devoted to methods of tracking the
measurement process all the way through the system to the system output.
Accuracy was addressed in terms of how well what was designed to be done was
actually done.  The second part of this sub-section will deal with how well the
system design serves the application.

       4.2.8.2 Summarized Data

       Summarization schemes are many and preclude a full discussion here.   The
auditor should define the methods used and comment on the appropriateness of
the method to the application of the summarized data.  There may be concurrent
summarizations such as a scalar wind speed, a resultant vector wind speed and
some kind of summarized wind direction.   The accuracy of the data system should
reflect estimated errors because of an inappropriate summarization program.

       A software analysis is required to be sure that the declared method of
summarization is in fact being accomplished by the computer program.   For
example,  is the scalar average wind direction avoiding the error of averaging a
circular range with a discontinuity?  If the average of winds ranging between
300  and 06O  turns out to be about 180°. this problem still exists.

-------
                                                              Section No.  4.2.9
                                                              Revision No.     0
                                                              Date:    17 Sep 89
                                                              Page:    1  of  3
4.2.9     REFERENCES
Acheson, D. T. ,1970:  Response of cup and propeller  ratios  and wind direction
    vanes to  turbulent wind fields. Meteor. Monogr.,  No.33,  Amer.  Meteor.  Soc.
    pp. 252-261.

Acheson, D. T.,1988:  Comments on "Anemometer Performance Determined  by ASTM
    Methods." J. Atmos. Oceanic Techno 1., 5 ,pp. 381-382.

ASTM. 1984:  Standard Practice for DETERMINING THE OPERATIONAL COMPARABILITY OF
    METEOROLOGICAL MEASUREMENTS. D 4430-84. Amer. Soc.  for  Testing and
    Materials, Philadelphia, PA 19103.

ASTM.1985aj   Standard Test Method for DETERMINING THE PERFORMANCE OF A CUP
    ANEMOMETER OR PROPELLER ANEMOMETER  (Draft 6 of D22.ll).  Amer.  Soc.  for
    Testing and Materials. Philadelphia, PA 19103.

ASTM.1985b:   Standard Test Method for DETERMINING THE DYNAMIC PERFORMANCE OF A
    WIND VANE (Draft 8 of D22.ll)  Amer. Soc. for Testing and Materials,
    Philadelphia. PA 19103.

Blackadar, A. F..1985:  Almanac for a weather station.  Heldref Publications,
    Washington. DC 20016.

Baynton. H. W. .1976:  Errors in wind run estimates from rotational anemometers.
    Bull.  Aater. Meteor. Soc., 57, 1127-1130.

Box. G. E. P.. W. G. Hunter and J. S. Hunter, 1978:   Statistics for
    experimenters. John Wiley & Sons, ISBN 0-471-09315-7.

EPA, 1987a:   Ambient Monitoring Guidelines for Prevention of Significant
    Deterioration (PSD). EPA-450/4-87-007, Office of Air Quality Planning and
    Standards. Res. Triangle Park. NC 27711.

EPA. 1987b:   On-Site Meteorological Program Guidance for Regulatory Modeling
    Applications, EPA-450/4-87-013, Office of Air Quality Planning and
    Standards. Res. Triangle Park, NC 27711.

Finkelstein,  P. L..1981:  Measuring the  dynamic performance  of wind vanes.
    J.  Appl.  Meteor., 20. pp. 588-594.

Finkelstein.  P. L..J. C. Kaimal. J. E.Gaynor, M. E.Graves and T.  J.Lockhart,
    1986:   Comparison of Wind Monitoring Systems. Part  I:   In-Situ Sensors.
    J.  Atmos. and Oceanic Techno I. , 3, pp. 583-593.

Finkelstein.  P. L.. J. C. Kaimal, J. E.Gaynor, M. E.Graves and T.  J.  Lockhart.
    1986:   Comparison of Wind Monitoring Systems. Part  II:   Doppler Sodars.
    J.  Atmos. and Oceanic Techno 1. , 3, pp. 594-604.

Fritschen. L. J.  and L. W. Gay, 1979:  Environmental  instrumentation.
    Springer-Verlag, N.Y. ISBN 0-07-033175-8.

-------
                                                              Section No. 4.2.9
                                                              Revision No.    0
                                                              Date:   17 Sep 89
                                                              Page:   2  of  3


Gill.  G.  C. , 1967:  On  the dynamic response of meteorological  sensors and
    recorders. Proceedings of the First Canadian Conference on
    Micrometeorology,  Part 1. Meteorological Service of  Canada,  Toronto.

Gill.  G.  C. ,1973:  The Hellcold Anemometer.  Atmosphere.  11, 4.  pp.  145-155.

Hayashi,  T. ,  1987:   Dynamic response of a anemometer.  J.  Atmos.  Oceanic
   Techno!.,  4. pp.  281-287.

Hoehne, W. E., 1973:  Standardizing Functional Tests. IEEE Transactions on
    Geoscience Electronics. Vol GE-11, No. 2, April.

Huschke.  R. E.,1970:   Glossary of Meteorology. Amer. Meteor.  Soc.,  Boston, MA
     02108

Kaiaal. J. C.  and  J. E. Gaynor, 1983:  The Boulder Atmospheric Observatory,
    J. Appl.  Meteor.,  22, pp. 863-880.

Kaimal, J. C. , J.  E. Gaynor,  P. L. Finkelstein, M. E.  Graves, and
    T. J. Lockhart,  1984:  A field comparison of in situ meteorological
    sensors.  NOAA/BAO  Report No. Six.

Kondo, J. , G.  Naito, and Y. Fujinawa, 1971:  Response of Cop  Anemometer  in
    Turbulence. J. Meteor. Soc. of Japan, 49, pp.63-74.

Lockhart, T.  J. ,  1977:  Evaluation of rotational anemometer errors. Bull.  Amer.
    Meteor. Soc.,  58,  pp. 962-964.

Lockhart. T.  J. ,  1978:  A field calibration strategy for rotating anemometers
    and wind  vanes.  Proceedings of the 4th Symposium on Meteorological
    Observations  and Instrumentation, Denver CO, April 10-14. pp.  57-60.

Lockhart, T.  J. ,  1985a:  Some cup anemometer  testing methods. J. Atmos.  Oceanic
    Techno1., 2,  pp.  680-683.

Lockhart, T.  J. ,  1985b:  Wind-Measurement Calibration. Bull.  Amer.  Meteor.
    Soc. , 66. p. 1545.

Lockhart, T.  J. ,  1987:  Performance of an anemometer determined by the  ASTM
    method. J. Atmos.  Oceanic Technol. , 4, pp. 160-169.

Lockhart, T  .J.,  1989  Accuracy of the collocated transfer standard method for
    wind  instrument  auditing. J. Atmos. Oceanic Technol. , 6,  pp. 715-723.  6/88)

MacCready. P. B..Jr. and H.  R. Jex,1964:  Response characteristics and
    meteorological utilization of propeller and vane wind sensors.  J.  Appl.
    Heteor..  3,  pp.  182-193.

MacCready. P. B.,Jr..1965:   Dynamic response  characteristics  of meteorological
    sensors.  Bull. Amer. Meteor. Soc., 46, 533-538.

MacCready. P. B. Jr..1966:  Wind speed measurements  in turbulence.  J.  Appl.
    Meteor.,  5,  pp.  219-225.

-------
                                                             Section No.  4.2.9
                                                             Revision  No.     0
                                                             Date:   17 Sep 89
                                                             Page:   3 of   3


Middleton,  W. E. K. and A. F. Spilhaus,  1953:  Meteorological  Instruments.
    University of Toronto Press.

Natrella,  M. G.,1966:  Experimental Statistics. National  Bureau of  Standards
    Handbook 91.

Snow, J. T., D. E. Lund, M. D.Conner, S. B. Harley  and C.  B. Pedigo.1989:   On
    the dynamic response of a wind measuring system. J. Atmos.  Oceanic
    Techno1.. 6, pp. 140-146.

Stearns, C. R.,1985:  Wind-Measurement Calibration, Response.  Bull.  Amer.
    Meteor. Soc.. 66, p.1545.

Sutton, 0.  G.,  1955:  Atmospheric Turbulence.

Turner, D.  B..1986:  Comparison of three methods  for  calculating the standard
    deviation of the wind direction. J.  Climate Appl.  Meteor.,  25,  pp. 703-707.

Wieringa.  J..  1967:  Evaluation and design of wind vanes.  J.  Appl.  Meteor.
    6. pp.  1114-1122.

-------
                                                            Section No. 4.3.0
                                                            Revision No.    0
                                                            Date:   17 Sep 89
                                                            Page:    1  of  3

                                 Section 4.3
               QA FOR TEMPERATURE AND TEMPERATURE GRADIENTS (AT)
                                   OUTLINE

Section                                                    Pages  Rev.  Date
4.3.0  OUTLINE AND SUMMARY                                  3     0     9/89

4.3.1  TYPES OF INSTRUMENTS                                 4     0     9/89
       4.3.1.1  TRANSDUCERS
                4.3.1.1.1  Thermistors
                4.3.1.1.2  Wire Bobbins
                4.3.1.1.3  Mercury-in-glass Thermometers
                4.3.1.1.4 Thermocouples
       4.3.1.2  SIGNAL CONDITIONING
       4.3.1.3  ASPIRATED RADIATION SHIELDS

4.3.2  SPECIFICATIONS                                       4     0     9/89


       4.3.2.1  DELTA TEMPERATURE (AT)
       4.3.2.2  TEMPERATURE

4.3.3  ACCEPTANCE TESTING                                   2     0     9/89

4.3.4  INSTALLATION                                          1     0     9/89

4.3.5  CALIBRATION                                          9     0     9/89

       4.3.5.1  EQUIPMENT AND METHODS
       4.3.5.2  DELTA TEMPERATURE (AT)
       4.3.5.3  TEMPERATURE
       4.3.5.4  CALIBRATION REPORT EXAMPLE

4.3.5  OPERATIONS. MAINTENANCE AND QUALITY CONTROL           3     0     9/89

       4.3.6.1  OPERATIONS
       4.3.6.2  MAINTENANCE
       4.3.6.3  QUALITY CONTROL

4.3.7  PERFORMANCE AUDIT METHODS                            8     0     9/89

       4.3.7.1  GENERAL CONSIDERATIONS
                4.3.7.1.1  Who
                4.3.7.1.2  What
                4.3.7.1.3  Where
                4.3.7.1.4  When
                4.3.7.1.5  How

       4.3.7.2  DELTA TEMPERATURE (AT)
                4.3.7.2.1  Sensor Control Method
                4.3.7.2.2  CTS Method

-------
                                                            Section No. 4.3.0
                                                            Revision No.    0
                                                            Date:   17 Sep 89
                                                            Page:    2  of  3
Section                                                   Pages  Rev.  Date
       4.3.7.3  TEMPERATURE
                4.3.7.3.1  Sensor Control
                4.3.7.3.2  CIS Method

4.3.8  ESTIMATING ACCURACY AND PRECISION                     1      0     9/89

       4.3.8.1  MEASUREMENTS
       4.3.8.2  SUMMARIZED DATA

4.3.9   REFERENCES                                           1      0     9/89

-------
                                                             Section No.  4.3.0
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    3  of  3

                       QUALITY ASSURANCE FOR TEMPERATURE
                         AND TEMPERATURE GRADIENT (AT)
4.3.0  SUMMARY

    The measurement of temperature Is standardized in great detail by all those
organizations interested in such procedures, ASTM, ISA (Instrument Society of
America),  and IMS  (Temperature Measurement Society).   The problem with
meteorological applications is that the free air temperature is required.  This
means the transducer needs to be exposed to the atmosphere which is in turn
exposed to the sun about half the time and to the very cold outer space the
other half.  The atmosphere is conditionally transparent to heat sources (the
sun) and sinks (outer space) so shielding must accommodate a wide range of
radiative conditions.  Wind also influences the temperature shield.  It
transfers beat to  and from the shield in a variable way as a function of wind
speed.  Most effective shields use forced aspiration to expose the transducer
to nearly unmodified outside air.  Wind speed may also play a variable role in
the performance of the aspiration system.
    This section concentrates on the meteorological applications of air
temperature aeasurement and  the differential temperature measurements which are
interpreted as  temperature gradients and applied as a measure of vertical
stability.

    Since the application of the measurement should define the accuracy needed,
both the relatively course air temperature and the relatively fine temperature
difference Measurements will be considered.

-------
                                                             Section No. 4.3.1
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    1  of  4
4.3.1  TYPES OF INSTRUMENTS
    There are several materials and structures which change in some way as a
function of  temperature.  General books on meteorological instruments such as
Mason and Moses  (1984), Middle ton and Spilhaus (1953) and particularly Brock
and Nicolaidis  (1984) will provide details on a variety of these sensors.  From
the standpoint of quality assurance, a few basic principles and a few standard
types will represent the vast majority of instruments in use for air quality
applications.

    The measurement of  temperature for air quality applications is generally
thought of as either air temperature, T, or a difference between two
temperature  measurements, AT.  The application of these different measurements
require different specifications and auditing methods.  The two types of
temperature  measurement will be treated in later sections as separate
measurements.

    Temperature  instruments are made up of three important parts.  The
transducer is the device which changes its electric value as a function of the
temperature  of  the transducer element.  The signal conditioner and cables
convert the  electric value to a recordable output, usually volts.  The
aspirated radiation shield is the mounting structure which holds the transducer
in the atmosphere where the temperature is to be monitored.  Each of these
three parts  will be discussed separately since there are various combinations
possible.

       4.3.1.1   Transducers

       Consider  the transducer as the part containing the sensing element.   In
most cases,  the  sensing element is  the transducer in air quality monitoring
applications.   The element is usually a thermistor (or thermistor network) or a
winding of fine  wire on an insulated bobbin.  It could also be a thermocouple
or a circuit element like an integrated circuit  (see Cole, 1978).  The elements
are usually  encased in  a protective capsule and  sealed.  From an operational
standpoint it only matters how the  sensor reacts to the temperature inside the
aspirated radiation shield.

                 4.3.1.1.1  Thermistors

                 The thermistor is an electronic  semiconductor made from certain
metallic oxides, such as nickel, manganese, iron, cobalt, copper, magnesium,
titanium and other metals.   It is a nonlinear element.  One common supplier
(Yellow Springs  Instrument Co. [YSI]) sells both the standard thermistor and
the "linear" thermistor.  Table 4.3.1.1 shows a  typical negative thermal
response curves  of raw  thermistors  and the nearly linear response of the
network thermistor.  Also shown Is  the positive  response of two platinum RTD
(resistance  temperature detector or resistance thermal device) for contrast.
Notice the large average change per °C with the YSI bead thermistor between
10 C and 20°C  (222O) as compared to the network thermistor (1260) or the 1000
RTD (0.4O) or  the  10000 RTD  (3.8O).  The raw bead thermistors are included
because, in  the  future, microprocessor-based data systems can handle nonlinear
transducers  as  easily as linearized ones.  The "linearized" YSI has a small
oscillating  error  of about ± 0.1°C with a wave length of about 40°C.  The
Impact of this  error on temperature difference systems is small.  At the

-------
                                                             Section No.  4.3. 1
                                                             Revision No.     0
                                                             Date:   17 Sep 89
                                                             Page:    2  of  4
                Table 4.3.1.1 - Sensor Resistance vs.Temperature


T
(°C)
-10

0

10

20

30

40

50




Thermistors


44031
(ohms) (An)
16600
-684.0
9796
-382.5
5971
-222.3
3748
-133.1
2417
-81.9
1598
-51.7
1081
44203 network
(ohms) (An)
13438
-127.9
12159
-125.3
10906
-126.1
9645
-128.6
8359
-128.8
7072
-123.6
5836
Platinum RTDs
____t_TKir~n___—

5-100 Pt
(ohms) (An)
96.09
0.391
100.00
0.390
103.90
0.389
107.79
0.388
111.67
0.387
115.54
0.385
119.39
	 HY-CAL 	
looon Pt
(ohms) (An)
961.84
3.816
1000.00
3.804
1038.04
3.792
1075.96
3.780
1113.76
3.768
1151.44
3.756
1189.00
(An) is ohms per degree C for the 10 degree range
VCT _______ V<

allow Springs, OH 45387
MINCO 	 7300 Commerce Lane, Minneapolis, MN 55432
HI-CAL 	 9650 Telstar Ave. El Monte, CA 91731-3093
steepest slope  it  is  0.025 C per Degree  difference.  A AT would need to be ±
4°C before the  error  reaches 0.1°C,  at which  point  the error  is moot with
respect to application.

                 The big  advantage to thermistors  is the  relatively  large
resistance of the  element with respect  to the resistance of the signal cable.
Lockhart and Gannon,  1978, pointed out  the fact that it  would take  a 12.5n
difference in cable resistance to two sensors of  a  AT pair to cause a 0.1 °C
error (bias) in the AT measurement.   Most signal  conditioning circuits have the
capability of adjustment to eliminate such a  bias.   Another little  known
advantage is stability.   A several year  stability test conducted at NBS showed
thermistors to  be  extremely stable.   This is  contrary to early experience which
suggested that  thermistors often failed  with  a shift in  the transfer function
(ohms vs. temperature) gaining them a reputation  of instability.  Better
packaging designs  and better handling practices have eliminated many of these
problems.

                 4.2.1.1.2  Wire bobbins

                 The resistance of a wire changes  with temperature.  If a long
piece of fine wire can be handled in some way.  it can be used as a  temperature
element.  Winding  the wire on a non-conductive bobbin is the  traditional method
of handling.  When the bobbin is large  and open,  a  very  fast  response sensor is
made.  When the bobbin is potted in a stainless steel Jacket,  a more
traditional slower response sensor is made.   The  transducer in both cases is a
length of wire.  Different metals have  different  temperature  coefficients.  A

-------
                                                             Section No.  4.3.1
                                                             Revision No.    0
                                                             Date:    17 Sep 89
                                                             Page:     3  of  4


40 gauge (0.08 nun diameter) wire at 20°C made of annealed copper has a
resistance of 3.40/m.   If It were made of German silver the resistance would be
65.9Q/B.  Platinum and iron are each 20.0 Q/m while nickel is 15.6 Q/m.
Manufacturers, choosing for stability, ease of handling and cost for a suitable
resistance, have settled on a few materials.  Platinum (1000) is the most
common for meteorological applications.  Nickel-iron is another common wire
providing a higher resistance at a lower cost with good stability.
                                                              o.
                Because of the small resistance change for a 1 C temperature
change (0.4ft), the transducer resistance must be measured with both high
resolution and attention to cable resistance.  Three and four wire bridge
circuits are commonly used, the latter being best for handling long cables.   It
will be shown in the sections on calibration and auditing that many of the
details of the transducer and circuits need not be known.  Only the system
performance is important.

                4.3.1.1.3  Mercury-in-glass thermometers

                These thermometers are not transducers, but they are commonlye
used for calibrations.  Some styles have enough resolution to be read to 0.02 C
with some care.  The ASTM series of Precision thermometers are examples of
these.  They are 37.9 cm (15 inches) long and breakable  (and expensive - $50  to
$80 each).  They are also calibrated for total immersion.  The ASTM 62C has a
range from -38 to +32°C and scale divisions of 0.1.  The ASTM 63C has a range
of -8 to +32°C and also has scale divisions of 0.1, but  they are farther apart
making interpolation more accurate.  If higher temperatures are needed, the
ASTM 64C has a range of 25 to 55°C.

                4.3.1.1.4  Thermocouple systems
                The thermocouple operates on the principle that when  two
different metals are joined, a  small voltage with a temperature- dependent
magnitude is generated.  By comparing  this voltage to  the voltage generated by
a  second  thermocouple in a thermally stable  environment  of known temperature,
the  temperature of the first thermocouple can be determined.  Because of
complex circuitry and problems  with conductors,  thermocouple systems  are no
longer popular  transducers for  meteorological monitoring.

                Thermocouple pairs are well  suited for differential  temperature
measurement.  They provide the  same voltage  for  any size wire which makes  them
ideal for miniature fast response applications.

       4.3.1.2  Signal Conditioning
       There  are a multitude of circuits which will measure  resistance.
Usually  the  transducer and the  signal  conditioner are  purchased  as a  system,
complete  with interconnecting cables.  This  is advisable since  the  range of
resistance vs.  temperature  is quite  large.   Signal conditioning  circuits may  be
adjusted  to  conform  to  Individual  transducers or transducer  pairs.   They may
also be  adjusted  to  a generic or theoretical curve or  transfer  function.   It  is
 important to  understand  the  function  of  the  signal conditioner  and  to treat  it
as a part of  a  system along  with the  transducers and  the cables.

-------
                                                                Section No.  4.3.1
                                                                Revision No.     0
                                                                Date:    17 Sep 89
                                                                Page:     4  of  4
       4.3.1.3  Aspirated  Radiation Shields
       There are many kinds of shields,  as Figure 4.3.1.1  depicts.   Most of the
error in measuring air temperature comes from the shield.   It is also true that
the magnitude of the error  is largely unknown.  In Section 4.3.2
SPECIFICATIONS, there is  a  detailed discussion about accuracy of shield
performance.

    The measurement of the  temperature of the free atmosphere at the point
where the  shield Intake is  located is the goal.  A shield  protects the sensor
from radiation and provides the mechanical mount for the sensor on a tower or
mast.  If  the shield is not aspirated, or designed for  effective natural
ventilation,  it may become  little more than a larger sensor case resulting in
the same radiation errors as an unshielded sensor but with a longer time
constant.   To avoid this  difficulty, it is necessary to draw the air into the
shield in  such a way that it is not modified by the shield temperature but will
cone to equilibrium with  the transducer at some average time.

       Forced aspiration  is the only way to minimize radiation error for all
conditions.   A fan draws  air in past the transducer at  a speed suitable for
minimum error.  Forced aspiration can be designed to provide a flow in the
right direction under all ambient wind conditions.  Insufficient pressure drop
in the fan during strong  winds may allow reverse flow to occur transporting the
heat from  the fan back to the transducer (Lockhart, 1975).
                      COW PorelM Pie Plote Shield
                    J
         A£S Stevenson Screen
          A£S PetuHelPlme
                                          Teledfne Atpireted
                                           Aediition Shi fid
                                                         ACS Atpireted
                                                        Stevenson Screen
                           JL
                          Imreeh TheOer Shield
        Kohl Setl-Afpireung Shield
              TT
             CaiNeturelrf
            Ventileted Shield

                                    Cbmet Meier Asptrued Shield
                                                        Modified Kemot Shield
ACS Merme Shield
                     ACS Duel Atpireted Pfnfmmeter

           Cirvod Plete Sr*old
                                            EG »G Dew Point
                                           Hygrometer Shield
                             Gill Atpireied Shield
 Figure 4.3.1.1 Examples of various radiation  shields
                 (McKay and HcTaggart-Covan.  1977)

-------
                                                             Section No.  4.3.2
                                                             Revision No.    0
                                                             Date:    17 Sep 89
                                                             Page:     1  of  4
4.3.2  SPECIFICATIONS
    The purpose of defining specifications is to give unambiguous meaning to
the terms  used by all those who are concerned that the instruments and systems
selected and operated will meet the needs of the application or project.   This
starts with procurement specifications and ends with supporting claims of data
quality. These specifications provide the basis for receiving inspection and
testing.

    Project and application requirements vary.   To make this handbook as
specific as possible, the examples used will be consistent with those presented
in the On-Site Meteorological Program Guidance for Regulatory Modeling
Applications (EPA, 1987b).  The specifications for temperature are range and
accuracy.   The performance of the radiation shield is not defined by
specification.   There is an Implication that the accuracy requirements include
this error source, but if they do there is no way suggested to verify the
performance of the shield.

       4.3.2.1  Delta Temperature (AT)

       The only requirement in EPA (1987a) regarding the vertical temperature
difference is "Errors in measured temperature difference should not exceed
0.003 C/m."  This rate is based on a 0.15°C accuracy for a 50m separation.
The requirement came from a time when AT was traditionally measured between the
lower 10m  level and the upper 60m level on a tower.  If a shorter tower is
used, like a 44m tower, the separation between 10m and 44m. namely 33m, would
show smaller lapse rates and inversions.  If the same accuracy were to be
preserved  in measuring the equivalent or representative AT from the shorter
tower, the measurement accuracy had to be better. 0.1°C in this case.  Some
operators  went to even shorter towers and to assure an appropriately accurate
measurement system, the requirement was stated as a per meter error.


       The above requirement is impossible to meet with the new 10m towers
including  AT.  If the aspirated radiation shields are mounted at 2m and 9m (to
avoid interference with the 10m wind), the requirement is to not exceed an
error of 7 x 0.003 = 0.021°C.  This is an accuracy which is hard to prove, let
alone achieve.
       The dynamic range for a AT installation on a 60m tower might be from
-2 C to +15°C.   Convention for positive and negative AT is:

    (-) a lapse rate is the normal decrease of air temperature with
        height limited by the auto convection rate of 3.4°C/100 m or
        0.034 C/ m.   A lapse rate produces a negative AT.

    (+) an inversion is the inverted lapse rate or an increase of air
        temperature with height.  There is no limit for inversion strength.
        An inversion produces a positive AT.

The dynamic range between 2m and 9m is not much different than that between 10m
and 60m.   During the EPA-BAO experiment in 1982 (Lockhart,  1988), a pair of AT
sensors was mounted on tower 4 at 2m and 8.6m.  Each sensor  was a 100ft Rosemont
platinum RTD in a Young aspirated shield.   On tower 3,  also  at 2m and 8.6m, a
pair of MRI-YSI linearized thermistors mounted in MRI shields (patterned after
the Young shields) were operated.  Figures 4.3.2.1 and 4.3.2.2 show three days
of 20-minute average AT data and 2m temperature data from tower 3.   Also shown

-------
                                                         Section  No. 4.3.2
                                                         Revision No.    0
                                                         Date:    17 Sep 89
                                                         Page:     2  of  4
           KEY: 	 MRI AT
                 +  VRI-BAO
      -2
         0    6    12   18    0    6    12    18    0    6    12    18    0
                 9/7/82
                                     9/8/32
9/9/82
Figurv 4.3.2.1  HRI AT Data and AT Difference Data for  Three Days
      30 -
               6    12   18   0
                                  C    12   18    0    6    12   18    0
                 9/7/82             9/8/82             9/9/82
Figure 4.3.2.2  Air Temperature at 2 m from the HRI AT Pair

-------
                                                             Section No. 4.3.2
                                                             Revision No.    0
                                                             Date:    17 Sep 89
                                                             Page:     3  of  4


is the difference between the two 20-minute average AT measurements from the
two towers.   Note the dynamic range of -1.3°C to +4.5°C (per 6.6m)  and the
agreement  between the two different instrument systems.  For theethree days the
average  difference is -0.02°C with a standard deviation of 0.10°C.   The daily
temperature  range was about 17°C.

      The reason the dynamic range is so high for such a small separation
distance is  that the surface is a better radiation receiver and transmitter
than is  the  air immediately above it.   The surface is almost always hotter or
colder than  the air above it.  Convection and mechanical turbulent mixing drive
almost all of the heat flux between the surface and the air.  The closer the
sensors  are  to the surface the larger the temperature difference per meter of
separation.   The drawback is that, the closer the AT pair is to the surface the
more sensitive the differential measurement is to local surface conditions or
character.   A black top road will affect a 2-10m AT much more than it will
affect a 10-60m AT.  The lower sensor really drives the AT and a 2m temperature
will vary more than a 10m temperature.

      For these reasons, the suggested procurement specification in the
On-Slte  guide (EPA, 1987b) in 8.1.3 reads:
     "Range                              -5 to +15 degrees C.
      Relative accuracy (error)         sO.l degrees C."

      while calibrations and audits of both accuracy and relative accuracy are
usually  conducted in controlled environments, the measurement is made in the
atmosphere.   The greatest source of error is usually solar radiation.  Solar '
radiation shield specification is therefore an important part of the system
specification.  Motor aspirated radiation shields (and possibly naturally
ventilated shields) will satisfy the less critical temperature measurement.  It
is critical  that the same motor aspirated shield design be used for both
sensors  used to measure AT.  The expectation is that the errors from radiation
(likely  to exceed 0.2 degrees C) will zero out in the differential measurement.
A motor  aspirated radiation shield specification might read:

      •Radiation range                    -100 to 1300 W/m2
      Flow rate                          3 m/s or greater
      Radiation error                    <0.2 degree C."

      Data sheets from five manufacturers (listed alphabetically) specify
their aspirated radiation shields as follows:

1.  Climatronics TS-10       Under radiation intensities of 1100 W/m2
                             measurement errors due to radiation will not
                             exceed 0.1 C.  Aspiration rate 3 m/s at sensor
                             location.

2.  Met  One  076              Radiation error -less than 0.05°F (0.03°C)
                             under maximum solar radiation of
                             1.6 gm-cal/cm2/min (1100 W/m2).
                             Flow rate 500 ft/min (2.5 m/s).

3.  Qualimetrics 8150-A      Radiation error - 0.05°C during maximum aspiration
                             and full sun.  Air speed 360 ft./min.  (1.8 m/s).

-------
                                                             Section No. 4.3.2
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    4  of  4

4.  Teledyne  Geotech 327C    Shielding - Under  test  radiation flux density of
                              1100 W/mo errors caused by radiation are less
                              than 0.1°C.  Aspiration rate 6 m/s at sensor
                              location.

5.  R. M. Young 43408        Radiation error -  under radiation intensity of
                              1080 W/m ,  Ambient  temperature - 0.2°C RMS,
                              Delta T - 0.05°C RMS with identical shields
                              equally exposed.   Aspiration rate - 3 m/s.

       It is  difficult not to notice the similarity  among these very different
designs.  An  auditor would need a comparative field  test to find the relative
error from  solar radiation.   Such a test can be done using a AT sensor pair
with two  (or  more)  shields collocated at the same level, 'if the transducers
are well calibrated,  the relative temperature of the transducers can be known
to 0.02 C.  these transducers in the two aspirated radiation shields will
report the  relative performance of the shields  to the same relative accuracy.
The one which is coolest in the daytime and warmest  at night has the least
radiation error.   Several diurnal cycles with sunny  days and clear nights are
required.   Such a test series could identify the most efficient shield which
could become  a standard against which a relative error analysis of any shield
could be made by a CTS method.

       4.3.2.2  Temperature

       The  accuracy specification for temperature is suggested in
EPA (1987b) as

       "Range                            -40 to +60  degrees C.
       Accuracy (error)                 so.5 degrees C."

       Some applications such as "PSD" permits  without fog problems require an
accuracy of only 1  degree C.   For locations with winds generally above 1.5 m/s,
a well designed naturally aspirated shield can  provide 1 degree C. accuracy.
When the application requires an aspirated radiation shield, the shield
performance requirement should also be included In the specification.  It is
customary to  use the lower AT shield to aspirate both the temperature and half
the AT pair.   Some designs develop AT by subtracting one temperature
measurement from the other,  in which case there is only one sensor in the lower
shield.

-------
                                                             Section. No. 4.3.3
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    1  of  2
4.3.3  ACCEPTANCE TESTING
    There are two ranges of temperature to consider.  One Is the measurement
range and the other is the environmental operating range.  The two might be
similar for remote Installations without commercial power or air conditioned
shelters.  The operating temperature range of the signal conditioning circuits
for remote installations is a function of the radiation shielding of the
electronics and the heat generated by the circuit operation.  The electronics
•ay get both colder and hotter than the air temperature under these
circumstances.  The operating temperature range for a station with an air
conditioned shelter is much narrower than the measurement range.  The receiving
or acceptance .test design must consider these factors.  A conventional
temperature chamber is required to control the temperature of the electronics
while the sensors are controlled by a separate thermal environment.

    A test which demonstrates the operation of AT and T sensors connected to
their signal conditioners is recommended.  The test should Include at least two
temperatures.  Liquid bath or solid thermal mass devices are recommended to
avoid local gradients in the air.  It is necessary to remove the sensors from
the aspirated radiation shield and place them in a bath while they are still
connected to the signal conditioner.   This may prove difficult to do with some
designs.  Since calibrations and audits are likely to require the same
accessibility of the sensors in the field, it Is best to solve this potential
difficulty In the less hostile receiving laboratory space.  Perhaps a statement
of accessibility needs to be Included in the specification.

    Experience has demonstrated that thermal stress can cause sensor failure
when baths are used.  The sensors may seem to be sealed, but if a sensor is
submerged in a hot (40 C) bath for enough time to reach equilibrium and take a
series of measurements, and then submerged in a cold  (0*C) bath, the pressure
change inside the metal sensor cover may draw water into the element chamber.

    It is prudent to assume that the sensors are not hermetically sealed and
to protect them as much as possible.   Using solid thermal mass devices is one
way to avoid liquid from wicking or being drawn into  the sensor, but it is not
a total protection.  In the above example, room air may be drawn into the
cooling sensor as the pressure inside equalizes with ambient pressure.  The air
drawn in may be saturated at the new cold temperature or may condense some
water vapor.  Whether or not this is a problem depends on how the sensor is
made, but it is best to take what precautions are possible.  If complete
Immersion is necessary, wrap the sensors in plastic so no liquid can get to the
interface where the wires come out of the sensor.  Use partial immersion where
possible keeping the interface dry.  Keep temperature changes small and in the
order AMBIENT-»COL&-» AMBIENT-»HOT-»AMBIENT.

    Assume the receiving test will use two temperatures, ambient temperature
and an ice slurry.  Assume a water bath will be used for the ambient test.  If
one to three sensors are in the system, wrap them together along with a
thermometer using a rubber band.  Use a Thermos" bottle which has been filled
with water several hours earlier.  The key to accurate temperature measurements
with sensors of different time constants is in having a thermal mass with
minimal gradients.  The value of a Thermos bottle is  its long time constant.
It will tend to keep the temperature of Its contents constant, but all it can
do is cause the heating or cooling to be slow.   In time the contents will be at

-------
                                                             Section No. 4.3.3
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    2  of  2

the same temperature as  the surroundings,  if the  temperature of the
surroundings  Is constant,  like an air  conditioned room.  When this equilibrium
has been reached,  the water in the bottle  will be at  the same temperature
everywhere In the  bottle and stirring  is unnecessary.  Stirring suggests the
need to mix up parts of  different temperature.  It is better not  to have parts
of different  temperature and this can  be achieved by  reaching equilibrium with
a well insulated mass.

    Place the sensors and thermometer  into the Thermos.  Use a cork or some
cover to keep ambient air froa circulating over the top surface of the water
mass.  After  about 30 minutes,  assume  the  sensors and thermometer are in
equilibrium (they  should be, they went in  from the same equilibrium
temperature)  and start a series of measurements.   Take five measurements about
five ainutes  apart.  If  the measurements are constant rounded to  the nearest
tenth degree  C, average  the five readings  and use them to describe the
temperature with respect to the thermometer and AT with respect to zero
difference.   If the measurements are slowly Increasing with respect to the
slower thermometer, there is a self-heating error.  Any resistance element will
get warm when current flows through it.  It is expected that the  self-heating
will be snail and  the large thermal mass will carry the heat away without
detection.  It is  also possible that the elements are sampled and do not have
current flowing continuously.  If self heating is detected, or if you wish to
shorten the time to equilibrium, some  mixing of the water in the  Thermos may be
useful.
    After the response to ambient temperature has been recorded,  place the
assembly of sensors and  thermometer in a Thermos  bottle containing an ice
slurry.  The  ice should  be made with distilled water  and crushed  into pea sized
pieces and mixed with distilled water  until an easily penetrable  slurry  is
reached.  As  long  as ice is present at the bottom of  the Thermos, the
temperature of the slurry will be 0.0  ±0.1°C.  Within 15 minutes  to one  hour,
equilibrium should be reached.  A series of five  measurements taken five
minutes apart should be  recorded.  If  the  measurements support the assumption
of equilibrium, then the five readings are averaged and recorded  as the
temperature relative to  0.0 C and the  thermometer,  and AT with respect to zero
difference.

    Accurately reading  the meniscus of the thermometer requires two things.
One. is the ability to see the meniscus and the scale at the same time.
Magnification is helpful even for those with good eyes.  It makes the
interpolation between scale marks possible and accurate.  Secondly, the  eye
muse be perpendicular to the meniscus  to avoid parallax errors.   If a mirror is
held against  the back side of the thermometer and the center of the image of
the eye moved to the meniscus level when the scale is read, the perpendicular
requirement will be met.

-------
                                                             Section No. 4.3.4
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    1  of  1
4.3.4     INSTALLATION
     Each design of aspirated radiation shield has its own installation
requirements.  The manufacturer's manual must be used in addition to the
general guidance given here.

     The Installation location is chosen to represent the temperature relevant
to the application.  The height above ground is the first consideration.  If
the temperature is to be used for climatological purposes, a 2m height above a
grass surface will do.  If the temperature is to be used to describe the air
being drawn into a manifold for chemical analysis, the best location is the one
that represents what the manifold inlet "sees."  If a temperature gradient is
to be used to describe the stability of the surface layer, a representative
pair of heights is selected.  Siting is best done with the concurrence of the
person who will be using the data for analytical purposes, the person who will
judge the data to be valid, and the person who will accept the data and the
analysis on behalf of the regulating agency.  Siting by this committee approach
will benefit from any objective knowledge any member might have, but its strong
point is in the mutual understanding of the criteria which were used in making
the selection.

 The second consideration is bias from surrounding structures.  The temperature
that is measured Is that of the air which is drawn into the aspirator.  If
there is a prevailing wind, mount the radiation shield into the wind such that
the wind passes the shield before reaching the tower.  The farther the
aspirator inlet is from the tower, the smaller the angle segment which can
contain the tower heat and the more mixing with non-heated air by the time the
inlet is reached.  The distance out from the tower should be the maximum
allowed by the mounting hardware.  Special booms for temperature may be
necessary if the design does not provideefor siting the inlet at least one
tower diameter from the edge, and if 0.5 C accuracy is expected.

-------
                                                             Section No. 4.3.5
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    1  of  9
4.3.5  CALIBRATION
    The manufacturer's manual will give instruction for the adjustment of
signal conditioning circuits in response to some input specification.  Usually
what is required is a precision resistor,  either built in or to be supplied by
the calibrator using a decade box or equivalent.  Such procedures are assumed
to have been done and will not be discussed here.  The method recommended in
this handbook is independent of the manufacturer.  It requires only the generic
transfer function of the resistance element used, that is the ohms vs.
temperature relationship upon which the measurement system depends for accurate
performance.

       4.3.5.1  Calibration Equipment and Methods

       The handbook method requires three stable thermal mass assemblies with
temperatures known to about 0.1 °C.  The three masses may be one mass used three
times with sufficient time allowed for conditioning to a new temperature.
Sufficient effort must go into the determination of thermal stability and
accuracy of the temperature measuring system used for that determination in
order to defend the results of the calibration or audit.  The following will
describe one solution to this requirement.  This solution is not the only one
but the confidence in stability and accuracy produced by this solution is an
example of the documentation necessary to support claims of accuracy.

    The thermal mass design uses a solid aluminum cylinder, shown in Figure
4.3.5.1, chosen for high thermal conductivity.  The mass is supported in the
space inside a stainless steel insulated bottle  (see Figure 4.3.5.2) by a
lucite tripod on the bottom and by three low thermal conductivity stainless
steel screw spacers at the top.  The lid is modified to allow transducer cables
to go through.  The insulating bottle is positioned inside a 2-gallon insulated
container modified to allow cables to go through with the top in place as shown
in Figure 4.3.5.3.
Figure 4.3.5.1 Thermal Mass
Figure 4.3.5.2  Insulated bottle

-------
                                                             Section No.  4.3.5
                                                             Revision No.     0
                                                             Date:    17 Sep 89
                                                             Page:     2  of  9
      Figure 4.3.5.3  Three mass containers:  Hot,  ambient,  cold (right)

The volumes and masses involved with this device are listed  in Table 4.3.5.1.
The air space  between the inside of the 2-gallon container  and  the outside of
the Insulated bottle is filled with either a cylindrical structure with a
heater strip and thermostat "floated" on stainless steel spacers  (#3, hot), air
(#2, ambient) or ice (#1, cold).

            Table  4.3.5.1  - Details  of  a Solid  Mass Thermal  Device
Component

Aluminum cylinder (less holes)
Holes
Air inside quart bottle
Quart bottle
Air outside quart bottle
2-gallon container
Total for #2 - ambient
Quart bottle (see above)
Ice-water mixture
Air above ice-water
2-gallon container
Total for #1 - cold
Volume
(cm )
485
58
1.100
931
5,882
10,206
18,662
2,574
4,706
1,176
10,206
18,662
Mass
(g)
1,341
0.1
1.3
1,154
7.1
1,600
4,104
2,496.4
4,706
1.4
1,600
8,804
Typical Transducers:
Minco S28F36Y nickel-iron 2.4 7
Rosemont RMT 78-39-7 5 36

-------
                                                            Section No.  4.3.5
                                                            Revision No.     0
                                                            Date:   17 Sep 89
                                                            Page:    3  of  9

    The ratio of 1.341 grams of thermal mass to about 43 grams of sensors meets
 the design goal of a small sensor  thermal impact. The several layers of
 insulation and the minimization of thermal conductivity paths meets the long
 time constant goal for the thermal mass.

 A test was conducted to document the performance of the thermal mass assemblies
 and to show the time required to condition the hot and cold masses and the time
 constant of the ambient mass, see  Figure 4.3.5.4.  From the beginning of the
 test at a little after 10:00 a.m.,  when ice was put Into #1 and the heater was
 plugged in for #3, it took about 9 hours for the slope to be flat enough to be
 confident that the mass had a homogeneous temperature.  The maximum rate of
 change of the ambient mass was 0.01°C/min.  The stability of the measuring
 circuit is shown by the line with  triangle symbol.  The thermal conductivity of
 aluminum is 0.5 cal./sec through a plate one centimeter thick across an area of
 one square centimeter when the temperature difference is 1°C.  Steel is 0.1,
 rubber is 0.0005 and air is 0.00005 (Hodgman, 1955).  The aluminum cylinder
 exchanges heat with its environment (except for the trantducers being tested)
   O
              Thermal  Mass  Conditioning  Test
                         Meteorological Standards Institute
                12
                             Local  Daylight Time
Figure  4.3.5.4
Thermal  mass conditioning and response test.
                                   2
through 450  cm  of air. 2.4  cm2 of lucite and 0.1 cm' of steel.   The difference
in temperature between  the holes where the transducers are mounted can be
estimated  from the difference  in conductivity within the aluminum and within
the environment outside the  aluminum (99.5%eair). or 10,000:1.   If the mean
temperature  of the cylinder  changes at O.Ol'c per minute, the gradient change
between hole temperatures could be 0.000001°C per minute.

-------
                                                             Section No.  4.3.5
                                                             Revision No.     0
                                                             Date:   17 Sep 89
                                                             Page:    4  of  9


    Having created a stable  environment,  there needs to be an accurate method
of measuring both the relative and absolute temperatures of the thermal masses.
A three transducer data system was designed for this purpose.  The report of
its calibration is shown at  the end  of  this section in 4.3.5.4.

       4.3.5.2  Delta Temperature (AT)

       Calibration of a AT system involves two parts.  One is the matching of
the transducers at zero difference and  the other  is the gain of the signal
conditioner for a known difference in temperature of element resistance.

       The first part involves placing  the AT transducers together in a series
of stable thermal masses.   A stable  thermal mass  is any mass which is at least
25 times the mass of  the sensors being  conditioned by the thermal mass and
which has thermal gradientsoof less  than O.Ol'c/cm throughout the mass.  Start
at cold, somewhere  in the  0°C to 5°C range, and record the system AT output
after stability has been reached in  30  to 60  minutes.  Take  readings about five
minutes apart.  When  the readings stabilize,  average the  last five.  Assume  the
output reports a difference of -0.02°C.  Then move the two transducers to the
thermal mass at ambient  temperature, somewhere  in the 15  C to 25 C range.  When
stability has nearly  been  reached,  start taking readings  about five minutes
apart.  When the  readings  have stabilized,  average the last  five  readings..
Assume the output  reports  a difference  of +0.03 C.  Finally, move the two
transducers to  the  hot  thermal mass, somewhere  in the 35  C to 40  C range, and
wait 30 to 6O minutes for  stability to  be reached.  Take  readings about five
minutes apart.  Assuming they are -0.09, -0.07, -0.05, -0.04. and -0.03,
stability has not  been  reached.  After  another  30 to 60 minutes  take another
series of measurements  about five minutes apart.  Now they are -0.01, +0.01,
0.00. 0.00 and  +0.01.   Stability has been reached and the average of the  last
five readings.  0.00,  is recorded.

       This  test  has confirmed  that the  two sensors  are matched  to each other
and to  the generic transfer function with which the  signal conditioning
circuits  have  been set.    It may be  that  the matching was  done  in the circuitry.
 It does not  matter.   It has been shown that the transducers  and  their  parts of
 the circuitry  agree with eachoother at three different  temperatures.If
agreement is not  within +0.05°C of  the true value of 0.00 C, look to  the  manual
or the manufacturer for guidance in correcting the problem.   The AT system
 should start off with agreement in  controlled conditions  of  much better than
0.1 °C  if  the atmospheric measurements  are to approach  that accuracy.   The
methods described here for  building a  stable thermal environment and  sampling
 the outputs  for zero difference are only an example  which works.   The  only
 important criteria is that  it  be documented  in terms of  stability,  whatever
 method is used.

        The second part of the AT calibration sets to tests  the gain of ,the
 difference amplifier.  Pick a common temperature for the site and substitute a
 fixed resistor for one transducer,  arbitrarily choose  the lower one.   Assume
 the transducer is a  100Q  platinum type (see  Table 4.3.1.1)  and your resister is
 108 ±10.   Substitute a precision decade  box  for  the upper transducer^   Adjust
 the decade box until the  AT output  is  the voltage equivalent of 0.00°C.   If the
 range is -5 to +15°C for  a  0  to 1 volt output, 0.00 C is 0.250 volts.   The
 output now reads 0.250 volts  and the decade  box  reads 107.96(2.  If 107.790

-------
                                                             Section No. 4.3.5
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    5  of  9

represents 20°C, and if 0.3890 represents a 1°C change (0.03890 for 0.1°C
change or O.01Q = 0.0257°C), the simulated temperature for both transducers is
(107.96 - 107.79 - 0.17; 17 x 0.0257 = 0.437) 20.437 ±0.026°C or between 20.41
and 20.46 C.

       If the decade box is changed to 108.35 (107.96 + 0.389), the upper
simulated transducer is now 1°C warmer than the lower simulated transducer.
The output should read +1°C or 0.300 volts (1.000 volts + 20 degrees = 0.050
volts/deg.; 0.250 + 0.050 = 0.300).  If zero and full scale are to be
challenged, set the decade box to 106.02 (107.96 - [5x0.389] = 106.02) for a
0.000 volt reading and set the decade box to 113.80 (107.96 +  [15x0.389] -
113.80) for a 1.000 volt reading.  Check the difference in box settings; 113.80
- 106.02 = 7.780; 7.78 + 0.389 « 20.00°C.

       Beware of rounding errors if enough resolution is not carried or
available on the decade box and the output.  Check the decade box with a good
ohm meter.  Errors in the tens wheel (when switching from 106 to 113) may be
larger than the smallest wheel.  If the tens wheel is only good to 1%. the
uncertainty is 0.100 or 0.26°C.

       4.3.5.3  Temperature

       The temperature calibration may be achieved concurrently with the AT
calibration.  Each thermal mass should have some accurate means of determining
temperature.-  While the actual temperature is not Important for the AT
calibration, it should be recorded on the calibration form.   If the temperature
transducer is not one of the AT pair, it can be placed in a thermal mass at the
same time the AT calibration is being done.

       It may be that the system does not have a AT measurement but it does
have a temperature measurement.  Remembering that the temperature accuracy
requirement is ±0.5 C, the temperature transducer can be challenged with a much
sinpler method.  Liquid baths in a pint or quart Insulated bottle with the
transducer and a good ASTM or equivalent thermometer mixing the bath together
will suffice.  Be sure stability has been reached before taking the readings.
Use care or parallax-avoiding devices when reading the thermometer.

       4.3.5.4  Calibration Report Example

       The following information is reproduced from a report prepared for
Meteorological Standards Institute to document the accuracy of temperature
instrumentation used on audits.

                4.3.5.4.1  Introduction

                During April 14 to May 8, 1988,  a calibration program was
conducted to verify the accuracy of three MINCO 604 ohm RTDs (resistance
theraal devices).  Three RTDs, Minco model S28F36Y labeled #1.  *2 and #3, were
originally calibrated in 1984 and have been in use for temperature and delta
tenperature auditing during the past four years.   The earlier calibration was a
relative calibration since the only accuracy of consequence to the application
was the inter- relationship of the three RTDs.   The current calibration is both
relative and absolute.

-------
                                                             Section No. 4.3.5
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    6  of  9
                 4.3.5.4.2  The measurement circuitry

                 The three RTDs are connected in series  to a battery powered
constant current source of 0.500 mA.   A fixed resistor  (668 ohms) of low
thermal sensitivity was also in the series loop as a reference  source  (REP).
The voltages  across each of the four resistors in the current loop were
connected  to  input channels of an ADC-1 data logger.  A NEC PC-8201A computer
controlled the ADC-1 and collected the data.   A program called  ADCT was used to
sample all the channels every 5 seconds and to record the average and  standard
deviation  for periods of time selected through the program.  Times of  10
minutes. 30 minutes and an hour were used at different  times during the
calibration.   The ADC-1 provides an output in tens of millivolts.  If  the
voltage across a 668 ohm resistor in a 0.0005 ampere current loop is 0.334
volts or 334  mV,  the ADC-1 will output 3340.

                 4.3.5.4.3  The conversion of Minco ohms to temperature

                 In the Minco Application Aid No.  7,  Table 14-604. the  nominal
values of  resistance for temperatures are given.   Nine  sets of  temperatures and
resistances in five degree steps from 0*C to 40°C were  used to  find a
mathematical  expression for converting resistance to °C.  A linear regression
was not satisfactory.  The quadratic solution to the regression analysis
predicted  the temperature at the nine points with an error of less than 0.01°C.
Table 4.3.5.2 shows the input pairs,  the predicted temperature  and error, and
the constants found and used.

        Table 4.3.5.2 - Relationship of Minco Resistance to Temperature
y
	 Minco Tab:
temperature
(°C)
0.000
5.000
10.000
15.000
20.000
25.000
30.000
35.000
40.000
X
.e 14-604 	
resistance
(ohms)
604.00
617.98
632.13
646.46
660.97
675.66
690.52
705.57
720.79
A
y
predicted
temperature
(°C)
0.006
4.998
9.994
14.996
20.001
25.005
30.005
35.004
39.994
y-y
error
(•ci
0.006
-0.002
-0.006
-0.004
0.001
0.005
0.005
0.004
-0.006
for y = a + bx + ex2 a = -269. 1531523
R => 1.00000009 b = 0.53213288
n = 9 c = -0.00014322177
                 A second step in the conversion requires  the constant current
to be exactly 0.5 mA so that the recorded voltage can be  converted  to
temperature.   The REF resistor is recorded as temperature using  the generic
conversion formula.   The difference between the recorded  value of REF and  the
correct value for 0.5 mA gives the correction.   To find the correct value  a
series of  measurements were made with the Fluke 8060A in  the current loop.  The

-------
                                                           Section No. 4.3.5
                                                           Revision No.    0
                                                           Date:   17 Sep 89
                                                           Page:    7  of  9

least squares straight line was  calculated for the data and both  the data and
the best-fit line are shown in Figure 4.3.5.5.  The REF value for 0.5 mA was
calculated and found to be 22.258 deg. C.
o
o
o
o

x
OT
<5
Q.
E
o
       5010-j
       5O09 -
       5O08 -
       5007 -
       5006 -
       5005 -
       5004 -
       5003 -
       5002 -
       5001 -
       5000 -
       4999 -
       4998 -
       4997 -
       4996 -
       4995 -
       4994 -
       4993 -
       4992 -
       4991 -
       4990 -
                   Circuit Current vs.  REF (#4)
                              Tests of May 3-5, 1988
            Regression  Analysis
            Y = 4497.33 + 22.584 X
                                       I
       21.9
                            22.1
      T	
  I  22.3
22.258
                                                            22.5
                           REF "Temperature" (deg.  C)
Figure 4.3.5.5  "Constant" current versus REF resistor "temperature"
                 The third step in converting the ADC voltage, V, to temper-
ature. T, comes from the  statement in the program ADCT which applies the
quadratic equation to the resistance, assuming 0.5 mA current.  That statement
is as follows:T = -269.15 + 0.10642V - 5.728E-06V2.

                4.3.5.4.4 Measurement of true temperature

                Two methods were used to find the true temperature.   The first
was the use of an ice slurry for 0°C.  The second was the use of an ASTM 63F
nercury-in-glass thermometer.  The Princo Instruments Factory Certificate of
Accuracy Tolerances for s/n 245453, scale range 18"F to 89°F with divisions
every 0.2 F.  states ±0.2°F or one division.   The thermometer was read with an
optical magnifier with anti-parallax targets  to the nearest 0.05°F or about
0.03 C.  In the relative  sense, the temperature should be accurate to 0.1°C and
in the absolute sense to  0.2°C.

-------
                                                             Section No.  4.3.5
                                                             Revision No.     0
                                                             Date:    17 Sep 8?
                                                             Page:     8  of  9
                4.3.5.4.5  The test  facility

               The three Minco RTDs  were  taped to  the mercury
bulb of the thermometer after all  four  devices were  threaded through a rubber
                          stopper.   The assembly was submerged  in  an ice slurry
                          or in water  in  a pint Thermos bottle  (see Figure
                          4.3.5.5).  The  data logger signaled when an average
                          was being  recorded and the thermometer was read at
                          the same time.  The average from  the  Mlnco RTDs
                          represents the  middle of the time period while the
                          thermometer was read at  the end of the time period.
                          A linear extrapolation of  the Minco RTD  data to the
                          time of  the reading of the thermometer provided
                          comparable data.

                          4.3.5.4.6  Results

                               There were 73 thermometer readings  over a nine
                          day period covering a temperature range  of 31.95°F to
                          89.05°F  (-0.03°C to 31.69 C).  The differences in
                          temperature between Mlnco sensors #1  and #3,
                          expressed  in  °C, and the thermometer  temperature,
                          expressed  in  °F, are shown in Figure  4.3.5.7.  Sensor
                          #2 was so  close to ttl that It was not plotted.   The
                          calibration correction curves for all three sensors
                          are shown  on  the figure.   The best fit lines from the
                          linear regression analyses are drawn  on  the figure.
                          The coefficients are listed in Table  4.3.5.3.

                MINCO  Thermometer  Calibration
                           Meteorological Standards Institute
Figure 4.3.5.6
 Calibration
      0.5
 O
 CD
      0.4 -
      0.3 -
 CD
 u

 v    0.2 -
 CD
 >4—
 V^

 b    0.1 -
                  KEY
              a   |3 (Regression Analysis

              0   #1 (Regression Analysis
            B
            f 3 Correction (-0.265° C)

            |2 Correction (-0.010°C)
            o
 a
 E
     -0.1 -
     -0.2
            |1 Correction (-0.002° C)
            	I	r—       —r-       —t—       —I—
         30                   50                   70
                  ASTM-63F  (245453)  Temperature (deg.  F)
Figure 4.3.5.7  Calibration of  Minco  sensors (RTDs)
                                                                        90

-------
                                                             Section No. 4.3.5
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    9  of  9
              Table 4.3.5.3
Regression output for Minco ttl and »3
versus mercury-in-glass thermometer

Constant (a)
Standard error of y
Coefficient (b)
Standard error of b
Number (n)
Average difference
Standard deviation
Minco #1
0.02262
0.02838
-0.00030705
0.00019593
73
0.002
0.028
Minco #3
0.29951
0.02534
-0.00041265
0.00017491
73
0.272
0.025
                The relative calibration, without consideration of an outside
measurement of temperature, covered a range of 0°C to 44 °C.  There were 254
averages recorded for each of the Minco RTDs.  The average and standard
deviation of the differences between RTDs is shown in Table 4.3.5.4.

                  Table 4.3.5.4  Relative difference analysis
Average difference
Standard deviation
	 Minco Sensors (RTDs) 	
#3 - #1
0.2634
0.0183
#3 - #2
0.2753
0.0165
*2 - #1
0.0119
                4.3.5.4.7  Conclusion

                Using^Minco ttl as a standard of comparison, and adjusting #1
for the bias of 0.002°C as shown in Table 4.3.5.3,  the following accuracies of
relative temperature are estimated.
           Minco *1
                 Minco #2
                 Minco #3
  *1 - 0.002
  #2 + 0.010
  *3 - 0.265
temperature ±0.05 C
temperature ±0.05°C
temperature ±0.05°C
Similar tests in 1984 yielded the following corrections:
                            #1  -i- 0.00
                            *2  + 0.03
                            #3  - 0.22
            temperature
            temperature
            temperature
showing reasonable stability over four years and a reasonable capability to
duplicate relative calibrations.

-------
                                                             Section No.  4.3.6
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    1   of  3

4.3.6  OPERATIONS. MAINTENANCE AND QC

       4.3.6.1  Operations


       From the standpoint of quality assurance, routine inspection of the
temperature and AT data will help find problems soon after they  occur.  Routine
inspection of the instrument system is also required.  The temperature
equipment is usually free from wear or change, except for the aspirated
radiation shield, which tends to attract homeless critters of all kinds.
Inspection of the temperature shield and transducer should be a  part of the
routine site visit and duly noted in the site log book.  If there is no routine
visit, a weekly or at the very least monthly visit is recommended.  Systems
usually have' some built in calibration feature which substitutes resisters for
the transducers to check or calibrate the signal conditioner.  The site visit
should include a temperature comparison with a simple hand-held  thermometer for
reasonableness.  Guard against radiation errors with the hand held thermometer
when working in direct sun light by keeping the thermometer in shade and  away
fro* body heat, including hot breath.  The long time constant of
mercury-in-glass thermometers (minutes) may make the inevitable  warming from
biasing heat difficult to observe.

       4.3.6.2  Maintenance

       Routine maintenance should include cleaning the aspirated radiation
shield and verifying its function.  This should not have to be done any more
frequently than the bearing tests of the wind equipment.  Since  it is necessary
to get the wind equipment down from the tower for the bearing tests, a cleaning
of the radiation shields at the same time would be prudent and economical.

       4.3.6.3  Quality Control

       The best way to keep a AT system operating "in control" is to use  a data
quality control inspection for this variable.   The QA Plan should supply  the
details of the AT inspection program.

       An important aspect of the inspection is the background of the
inspector.   Ideally,  an experienced QC meteorologist should be used.  Lacking
this resource,  a training program should be made available to the person  who
will routinely perform the data inspections.   The training will  point out the
nature of AT data as a function of wind speed, cloud cover,  and  time of day
(solar angle).

       Additionally,  training will point out that a AT value,  that is the
difference between two well calibrated and shielded transducers,  is Just  that.
It is not a gradient  measurement unless there is reason to believe that the air
between the two transducers is reacting normally to thermal  flux.  Cases have
been observed  where a 10m to 60m AT averaged in excess of the auto convection
rate for hours.   The  easy assumption is that there is an instrument error
because autoconvection rates cannot be exceeded for long periods of time.   The
often unspoken assumption is that the AT transducers are in  the  same boundary
layer and the  difference in temperature represents the stability condition of
the air.   If the site can produce shallow or transitory surface boundary
layers,  as can happen with land-water interface regions,  one transducer may be

-------
                                                             Section No. 4.3.6
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    2  of  3


in one layer and the other  transducer  in  the other  layer.  Then the AT value
represents nothing more  than single  samples in different layers and lapse rate
conclusions are invalid  but the data are  valid.

       Training will show the normal diurnal cycle  from a negative AT (lapse
rate) in the daytime to  a positive AT  (inversion) at night  (see Figure
4.3.2.1).  An understanding of the physical process will support  the data with
other observations of weather conditions.  The sun  heats the surface much more
than the air above it.   The air at the surface is warmed by the now warmer
surface.  The warmer air expands and rises and mixes with the air it passes.
This unstable convective process continues until the driving force, the surface
warmer than the air above it, is neutralized.  This can happen by either
changes in the radiational  heating of  the surface or by the effective cooling
of the surface through the heat removal process described above.  Considering
the strength of radiant  heating (sun angle and sky  cover) and the strength of
the mixing process (wind speed), the size of the lapse rate (-AT) can be
imagined.

       Conversely at night, the surface is cooled by long-wave radiant  loss to
the cold universe.  As the surface gets colder than the air above it that
surface air is cooled by conduction.  The cooled air is mixed by  mechanical
turbulence, caused by the wind flowing over surfaces elements, and slowly cools
the air from the surface up.  This very stable process results in the air
above, not yet cooled, being warmer  than  the air below or an inversion  (+AT).
The size of the inversion results from the amount of surface cooling (sky
cover) and the amount of turbulent mixing (wind speed and surface roughness).
Calm clear nights will have larger  inversions than  cloudy and windy nights.   If
the wind is too high, there may not  be an inversion at all.  All  the air may be
mixed so well that there is no measurable difference between the  two AT
transducers.

       To complicate  the picture, temperature measurements  also change  as
different air with different conditioning history  is blown  or advected  past the
measurement site.  Diurnal rhythms  can be seen as colder air flows down terrain
features, such as hills  and valleys, at night as a  consequence of the kind of
surface cooling described above. These vertically  stable flow fields often
become decoupled from each other.  It  is  possible for one transducer to be  in
one stream of stable air while the  other  is  in another stream.  When this
happens, the AT value may not represent a temperature gradient as much  as two
separate flow regimes, similar to the  boundary  layer example.

       The trained  inspector learns  to see  these processes  in the AT data and
to recognize physically  unusual or  unlikely data.   If the data QC inspector
looks at the data on a weekly schedule, problems will be uncovered shortly
after they occur,  thus avoiding long periods of data  loss.  One week is also
short enough  to allow memory of conditions  to be correlated with  the data.
Discrepancy reports originated by the  QC Inspector  can initiate the testing and
corrective action.  If necessary, by the instrument  operator.  If  At
measurements are  taken,  the purpose is usually  the  determination  of stability
categories or parameters for use in diffusion models.  This important variable
deserves careful attention and well  documented  evidence  for validity claims.
The accuracy of  the measurement is  achievable only  with  careful calibration.
The accuracy of  the data requires as much care  given  to  data  inspection between
calibrations.

-------
                                                            Section No.  4.3.6
                                                            Revision No.     0
                                                            Date:    17 Sep 89
                                                            Page:     3  of  3
       If only temperature is measured,  the accuracy requirements are less
difficult and the data QC inspection can be less rigorous.   If data are
collected,  however,  routine inspection is recommended and  a weekly period is
reasonable.   The inspection compares the temperatures to reasonable values for
the week and the nature of the temperature change to realistic patterns.

-------
                                                             Section No. 4.3.7
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    1  of  8

4.3.7  PERFORMANCE AUDIT METHODS

       4.3.7.1  General Considerations

       A performance audit is a measurement made by an independent method and
person of the accuracy and precision of the performance of the measurement
system.  To make this determination for temperaturemeasurements, knowledge of
the input conditions to the system sensors is required.  It is also necessary
to know what the system will do to these input conditions in producing an
output.  The output is simply the system output.  Given the inputs and the
transfer functions, the output can be predicted.  The difference between the
predicted output and the system output is the error of the system or its
accuracy.  The methodology starts with the ways of controlling and measuring
the sensor inputs or knowing the inputs in an uncontrolled environment.
Prediction of output from the controlled input requires knowledge of the
transfer function but not necessarily its validity.  Knowing the conditions of
the uncontrolled environment does not require knowledge of any Intermediate
steps such as transfer functions.  Temperature in is simply compared to
temperature out.  The method using the latter approach is called the Collocated
Transfer Standard (CTS) method.  The CTS method is seldom practical for AT.
The best performance audit uses both methods where appropriate.

       A performance audit must follow some written procedures.  Since the
procedures must be relevant to the design of the instrument or system being
audited, only general principles will be described below with some specific
examples.  The data from the audit should essentially fill out an audit form.
It is important, however, for the auditor to be sufficiently experienced to be
able to deviate from the procedure or the form when the pursuit of truth leads
away from the expected.

       The starting point of an audit form is the documentation of the who.
what, where, when, and how the audit values were acquired.

                4.3.7.1.1  Who

                The performance audit report form should contain a space to
identify the auditor.  The audit report which summarizes the audit findings
should report the names and affiliations of the operators of the system.

                4.3.7.1.2  What

                The form should contain a section to identify the instrument
being audited by manufacturer, model number, and serial number.
Sub-assemblies, such as transducer or RTD and aspirated radiation shield.
should be identified by number.  If they are not numbered, the operator should
be asked to mark them for Identification.

                The audit report should contain a list of all the equipment
provided and used by the auditor, including model and serial numbers and time
of last calibration, where relevant.

                4.3.7.1.3  Where

                The audit form should have a space to show the location of the

-------
                                                             Section No. 4.3.7
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    2  of  8


sensor on a  tower, .including height.   A sketch  is  useful  to-show  the relative
positions of the  sensing elements with respect  to  possible  biasing  influences,
such as the  tower, other sensors and buildings.

                4.3.7.1.4  When

                The  date or dates when the audit affected the  system operation
should be listed.  The time when the system or  a particular sensor  was  taken
"off-line" and put back "on-line" should be listed.   The  time,  or time  period,
when each data value was taken is vital for the comparison  with the system
output.  Implicit in this is the need for the time the  auditor uses to  be
correlated to the time the operator or the system  uses.   The auditor should
rely on the  National Bureau of Standards station WWV for  correct  time.  Battery
operated receivers,  such as the Radio Shack Time Cube,  are  generally available.

                4.3.7.1.5  How

                The  audit form should either contain a  copy of the  method used
or reference the  method number.  The audit report  should  contain  copies of the
audit method used.   The methods should be detailed enough to  identify each step
in the acquisition of the audit value and in the  conversion of the  value to
units compatible  with the system output.  See Figures 4.3.7.1.  4.3.7.2.  4.3.7.4
and 4.3.7.5  for examples of forms and methods.  These figures  are intended as
aids for writing  specific methods and drawing the  companion form.  These may be
reproduced  if they are relevant.

       4.3.7.2 Temperature Difference  (AT)

                4.3.7.2.1  Sensor control method

                 The  audit method should simulate  the most complete calibration
method  (see  4.3.5).   The first step  is  to condition the thermal mass  assemblies
to be used  to challenge the transducers to a zero difference  environment.   If
stirred baths are to be used, be prepared to give ample time  for equilibrium.
The amount  of time is not well predicted by the transducer  time constant.
Stability at the 0.01°C resolution scale is desirable to back up claims of
accuracy to 0.1 C:

                 The transducers must be removed from the tower along  with  their
cables.  At  some installations  it will  be difficult to  impossible to  remove  the
cables.  Substitute cables may be used  if care is taken to make sure  the
substitute  does not change the output more than 0.01°C.  This should  be
verified by using a fixed resistance representing ambient temperature as a
substitute  transducer.  Output  readings using both the  operational cable and
the substitute cable should be  recorded on the audit form.

                 The AT audit  usually requires much more elapsed  tiire  than
measurement time.   It is practical  to have equipment which allows the
transducers to slowly reach  their equilibrium state and to record this process
through the entire  system.  This way the elapsed  time can be used for the  more
labor  intensive audit variables.

-------
                                                         Section No. 4.3.7
                                                         Revision No.    0
                                                         Date:   17 Sep 89
                                                         Page:    3  of  8
         Temperature Difference - MSI  method  OLT006 (version  8/1/84)

    This method provides for a measure of  two temperature transducers
    Hith  electric output used in a differential   application,   often
    called delta T.   The audit equipment includes  three thermal  mass
    assemblies.  Each is a cylinder of aluminum (6.4 cm diameter by
    17 cm long, 1,341 g in 48S cc)  with holes for  different kinds of
    sensors.  This 485 cc mass is suspended in the lrlOO cc inside
    volume of a stainless steel Thermos bottle, which  is inside  the
    8500 cc inside volume of an 18,5OO cc  cylinder (Gott 2 gallon
    water cooler).  The cold system is filled (4,500 cc)  outside the
    Thermos bottle with a mixture of ice and  water. The ambient
    system is full of ambient air.   The hot system has a cylindrical
    frame spaced outside the Thermos, bottle with a 600 watt strip
    heater operating through a 100 F thermostat.   Conditioning for
    the hot and cold masses begins at least  12 hours before planned
    use.  Each mass has a Minco 604 ohm (at O C) nickel-iron RTD.
    Resistance is measured and 'converted to temperature using a
    quadratic expression of the Minco Table  14-606 (see appendix).
    Relative corrections from intercomparisons made with the three
    RTDs in the same mass are applied yeilding relative accuracies of
    better than 0.05 C.  Absolute accuracy is better than O.5 C.

    DLT006-A  This method challenges the delta T pair  at zero
    difference at three different temperatures. Place the pair  of
    sensors in the hot mass.  Record the mass temperature and the
    sensor temperatures  (or the difference if that is  all the signal
    conditioner provides) after about 40 minutes.   Take two more
    samples five to ten minutes apart to verify stability.  Move the
    pair of sensors to the ambient mass.  Note: if DLTOO6-B is to be
    used, leave the hot mass unplugged and open to the air.  Record
    data after 4O minutes and again in five  to ten minute steps.
    Move the pair of sensors to the cold mass and  record data after
    40 minutes and again in five to ten minute steps.

    DLT006-B  This method challenges the delta T pair  at a small
    temperature difference.  The two masses  to use are the ambient
    and the hot, after the hot has cooled  down from use in DLT006-A
    or has been conditioned by a short input  of heat.   The true
    difference in temperature between the  two masses should be
    oetween one and three degrees C.  Install one sensor in the
    ambient mass  (T 1) and one sensor in the "hot" mass  (T 2).  After
    4O minutes record the temperatures of  the two masses and the
    sensors or sensor difference if that is all that is available.
    Record the data again after five to ten  minutes.  The true
    difference will change slowly as the masses change temperature.
    Reverse the sensors and repeat the method waiting  4O minutes for
    the first data point.  If the time constant of the sensor is
    observed to be fast enough to assure stability in  less than 40
    minutes, a shorter period may be used.
Figure 4.3.7.1  Audit method for temperature difference

-------
                PERFORMANCE AUDIT REPORT by
     MEASUREMENT SYSTEM - Temperature difference
           System number                        —
           Sensor        	..
           Transducers   _.
           Locations
           Si gnal_condi t i oner
           Data channels    . .
     DATE 	/	/	 TIME of-f line
                            on line
      TRANSFER FUNCTION  
-------
                                                             Section No. 4.3.7
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    5  of  8

                The transducers are challenged with a known zero difference at
three teoperatures.  This shows how well the transducers are matched and how
well they follow the generic transfer function.  The acquisition of these three
data points nay take four hours.  Some Judgment is required to Identify the
point at which stability Is achieved.  One clue is the reversal of a
progression of data points.  If, for example, the five minute average AT values
are 0.09. 0.07, 0.06. 0.06, 0.05, 0.05, 0.06, 0.05, the last 0.06 can be taken
as a signal that equilibrium has been reached.  The reported value could be the
average of five stable points (0.05) or the final stable value.

                After the zero difference test has been completed, the gain of
the difference circuit can be challenged by using the substitute resistance
•ethod.   This method is described in 4.3.5.  It is possible to use thermal
•asses at different temperatures, but It is not recommended because of the
uncertainties of stability and four transducers measuring two temperatures to
an accuracy of at least 0.025°C.  Once the matching at zero difference has been
established, the gain is adequately verified by normal electronic circuit
procedures.

                It is not practical to mount a second pair of sensors and
aspirated radiation shields for collocated testing.  The interference problems
with the aspirators are hard to overcome.   The physical problems associated
with mounting parallel instruments are large compared to the value of the
•ethod.   Considerable, but much less, effort is required for the Sensor Control
•ethod which provides numbers with acceptable confidence in their accuracy.

       4.3. 7.3  Temperature

                4.3.7.3.1  Sensor control method

                Usually the temperature transducer. If it is different from one
of the AT pair, can be included in the thermal mass with the AT pair, or in
another thermal mass at the saae time as the AT pair is being tested.  Timing
for stability can include temperature with AT as though It were the same test.
The big difference, however, is that the temperature transducer output is
coapared to the thermal mass transducer output as the audit value.  A
calibration of the thermal mass transducer is the key to the claim for accuracy
given the challenged system.

                If there is no AT system being audited, a simpler method is
appropriate for the temperature system.  A two point check using an Ice slurry
and an ambient bath is acceptable.   Two insulated bottles (pint or quart size)
with a cork to support a calibrated thermometer or a calibrated electric
thermometer are required (see Fig.  4.3.5.6).Stability is easier to find since
the readings are only taken to the nearest 0.1°C.

       4.7.3.2  CTS Method

       It is both practical and recommended to use the CTS method for
teaperature audits.  The temperature transducer and its aspirated radiation
shield (or even naturally aspirated shield) is usually located at an easily
reached elevation.   A CTS such as the Assmann Psychrometer shown in Fig.
4.3.7.3 can be located near the temperature sensor.  It should be exposed so
the wind reaches the CTS without bias error from other structures.  If the wind

-------
                                                             Section No. 4.3.7
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    6  of  8
during the audit is passing through a tower to reach the temperature sensor,
the CTS should not be exposed to sample the same biased air temperature.  The
presumption is that the temperature sensor represents the air temperature.  Any
error from siting is a part of the measurement error.  The CTS should be
mounted to avoid all bias, if possible.

                The CTS method should be
used as an additional challenge to the
temperature system.   The two point Sensor
Control method is a challenge to the
transducer and signal conditioning circuit.
The CTS method is a challenge to the
radiation shield at one point on the range
scale.  A one point challenge of a
temperature system   provides no
information about other temperatures.  It
may be that the operator calibrated the
system at one point, perhaps the same
ambient temperature as exists during the
audit.  There could be a slope error which
causes large errors at near freezing.
Having an accurate temperature measurement
near freezing, accurate to 0.5 C, can be
valuable as it relates to other sensors,
such as wind.   If the wind vane does not
show any direction variation and the
temperature system reports 0.0 C, there is
a good chance that ice is on the direction
vane bearing assembly.  If the temperature
system reports 3 C,  with a 3 C error
because only one point was used in
calibration and audit, a different estimate
of the direction vane problem is necessary.
If audit or calibration records exist
showing the full range performance of the
temperature circuit, a one point spot check          Figure 4.3.7.3
with a CTS is useful.                              Assmann psychrometer

-------
                                                         Section No. 4.3.7
                                                         Revision No.    0
                                                         Date:   17 Sep .89
                                                         Page:    7  of  8
             Temperature  -  MSI method  TEM005  (version 8/1/84)

       This method  provides -for a comparison of a temperature transducer
       with electric  output to a calibrated transducer in a sionly
       changing thermal  mass at three different temperatures;  The audit
       equipment includes  three thermal mass assemblies.  Each is a
       cylinder of  aluminum  (6.4 cm diameter by 17 cm long, 1,341 g in
       485 cc)  with holes  -for different kinds of sensors.  This 485 cc
       mass is  suspended in the l,10O.cc  inside volume of a stainless
       steel Thermos  bottle, which is inside the 8500 cc inside volume
       o-f  an 18,500 cc cylinder (Gott 2 gallon water cooler).  The cold
       system is filled  <4.5OO cc) outside the Thermos bottle with a
       mixture  of ice and  water.  The ambient system is full of ambient
       air.  The hot  system has a cylindrical frame spaced outside the
       Thermos  bottle with a 600 watt strip heater operating through a
       1OO F thermostat.   Conditioning for the hot and cold masses
       begins at least 12  hours before planned use.  Each mass has a
       Minco 604 ohm  (at 0 C) nickel—iron RTO.  Resistance is measured
       and converted  to  temperature using a quadratic expression of the
       Minco Table  14—606  (see appendix).  Relative corrections from
       inter comparisons  made with the three RTDs in the same mass are
       applied  yeilding  relative accuracies of better than 0.05 C.
       Absolute accuracy is better than O.5 C.

       TEMOO5-A  Insert  the RTD being challenged in the cold mass.  Wait
       about 30 minutes  or until stable temperature is reached.  Record
       samples  of the RTD  temperature from the system output.  Record
       the resistance measurements of the mass RTD.

       Move the RTD being  challenged  to the ambient mass and repeat the
       above procedure.

       Move the RTD being  challenged  to the hot mass and repeat the
       above procedure.

       TEMOO5-B  Use  an  Assmann aspirated psychrometer mounted in the
       vacinity of  the shielded temperature sensor.  Wind the Assmann
       and let  it run five minutes.   Wind again and after an additional
       two minutes, begin  reading the mercury—in-glass thermometers.
       Use the  anti-parallax magnifiers.  Record the temperatures from
       the Assmann  and from the system taken at the same time.  Be sure
       the two  sensor systems are sampling from air which has not been
       biased by local mounting structures.
Figure 4.3.7.4  Audit Method for Temperature

-------
                                                        Section No.  4.3.7
                                                        Revision No.    0
                                                        Date:   17 Sep 89
                                                        Page:    8  of 8
               PERFORMANCE AUDIT REPORT by

    MEASUREMENT SYSTEM - Temperature
         System number        	
         Sensor                	
         Transducer    		
         Location      	
         Signal  conditioner
         Data channel     _.
    DATE  	/	/	 TIME o-f-f line
                   on line
    TRANSFER FUNCTION (volts per degree C)
                       Conversion -formula
    TEST  RESULTS

      TEMOO5A         cold
               mass   output   di-f.
    time  ohms degC'volt degC  degC
                                                                TEMOOS
                                                       test start
                             ambient
                             mass  output   di-f.
                   time ohms degC volt degC  degC
          average              	

                       hot
               mass   output   di-f.
     time ohms degC volt degC  degC
          average
       TEMOO5B
          time
    Assmann
 dry   wet  RH
degC  degC   7.
  output

volc   degC
difference

   degC
Figure 4.3.7.5  Audit Form for the Temperature Method

-------
                                                             Section No.  4.3.8
                                                             Revision No.     0
                                                             Date:    17 Sep 89
                                                             Page:     1  of  1
4.3.8  ESTIMATING ACCURACY AND PRECISION
    Section 4.1.5 contains a detailed discussion of methods of estimating
accuracy,  precision and bias using wind speed and wind direction as examples.

    Temperature transducers (as differentiated from temperature sensors which
•ay include the radiation shielding) may be calibrated or audited by exposing
thai to a controlled environment such as a wet or dry thermal mass.  The
temperature of the mass is known either by an installed calibrated transducer
or by collocating a calibrated transducer in the mass.  The collocation
alternative requires assurance that the mass is at a-homogeneous temperature.
The installed option requires assurance that the installed location is
representative of the homogeneous mass temperature.  When such a method is
used, traditional statistical or metrological methods may be used to estimate
accuracy.   The precision of the method is the standard deviation about the
series mean value of repeated measurements in the constant and controlled
environment.  Such methods are capable of achieving accuracies of 0.1 degree C
and precision of 0.05 degree C or less.

    When a collocated transfer standard (CTS) method  is used, considerable care
is required in stating the accuracy of the calibration or audit.  The different
exposures in the atmosphere of different transducers with different time
constants in different radiation shields puts a larger uncertainty on the
coapar-ison than is found with a control-led environment.  If, for example, a 2
•eter temperature instrument is compared to an As smarm thermometer mounted
nearby at the same height, the accuracy claim might be no better than the sum
of the two different instrument accuracies.  The accuracy of the method might
be 1 degree C if each instrument is capable of a measurement accuracy of 0.5
degree C.   There is reason to consider such a method  as a comparative
measurement rather than an audit or calibration.

     It is possible that a CTS method can have greater accuracy.  What is needed
is a body of data which sets the functional precision of the CTS method by
finding the best one can expect from collocated temperature instruments.  A
companion requirement is a body of data which compares different radiation
shields as a function of radiation intensity, wind speed and wind direction
relative to the orientation of the aspirator motor.

-------
                                                             Section No. 4.3.9
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:    1  of  1
4.3.9  REFERENCES
Brock.  F. V. and C. E. Nicolaldis. 1984:  Instructor's Handbook on
    Meteorological Instrumentation.  NCAR/TN-237+1A, National Center for
    Atmospheric Research.  Boulder, CO.

Cole. H.  L.. 1978:  Air temperature and differential temperature measurement
    using 1C temperature sensors.  Preprints of Fourth Symposium on Met.
    Observ. and Inst.. Amer. Meteor. Soc..  Boston, MA., pp. 25-30.

EPA, 1987a:  Ambient Monitoring Guidelines for Prevention of Significant
    Deterioration  (PSD). EPA-450/4-87-007.   Office of Air Quality Planning and
    Standards, Research Triangle Park, NC.

EPA. 1987b:  On-Site Meteorological Program Guidance for Regulatory Modeling
    Applications. EPA-450/4-87-013.  Office of Air Quality Planning and
    Standards, Research Triangle Park. NC.

Hodgman,  C. D. , editor in chief. 1955:  Handbook of chemistry and physics.
    Chemical Rubber Publishing Company, Cleveland, OH.

Lockhart, T. J., 1975:  Variable errors in operational data networks.
    Proceedings of the 3rd Symposium on Met. Observ. and Inst.. Amer. Meteor.
    Soc., Boston, MA., pp. 91-96.

Lockhart. T. J. and M. T. Gannon,  1978:  Accuracy and precision of field
    calibration of temperature difference systems.  Proceedings of the National
    Conf. on Qual. Assur. of Environ. Meas.. Denver, CO., Nov. 27-29.

Mason,  C. J. and H. Moses, 1984:   Atmospheric Science and Power Production.
   Darryl Randerson, Ed.. DOE/TIC-27601, pp. 103-109

McKay,  D. J. and J. D. McTaggert-Cowan. 1977:  An intercomparison of radiation
    shields for auto stations.  Vorld Meteorological Organization Publication
    No. 480. pp. 208-213.

-------
                                                             Section No. 4.4.0
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page     1  of  7


               QUALITY ASSURANCE FOR PRECIPITATION MEASUREMENTS

4.4.0  INTRODUCTION

    By definition, "The total amount of precipitation which reaches the ground
in a stated period is expressed as the depth to which it would cover a
horizontal projection of the earth's surface if there were no loss by
evaporation or run-off and if any part of the precipitation falling as snow or
ice were melted"  (WHO, 1971).  In any method of precipitation measurement, the
aim should be to obtain a sample that is representative of the fall in the
area.  At the outset, it should be recognized that the extrapolation of
precipitation amounts from a single location to represent an entire region Is
an assumption that is statistically questionable.   A network of stations with a
density suitable  to the investigation is preferable.

4.4.1  TYPES OF INSTRUMENTS
    Precipitation collectors are of two basic types:   nonrecording and
recording!

       4.4.1.1 Nonrecording Gages

       In its simplest form, a precipitation gage consists of a cylinder, such
as a can with straight sides, closed at one end and open at the other.  The
depth of the liquid in the can can be measured with a measuring stick
calibrated in subdivisions of centimeters or Inches (Figure 4.4.1).

    To obtain greater resolution,  as in the case of the standard 8-inch gage
•ade to NWS Specification No. 450.2301, the gage is constructed with a ratio of
10:1 between the area of the outside collector cylinder and the inside measuring
tube.  The funnel attached to the collector both directs the precipitation Into
the tube and minimizes evaporation loss.  Amounts in excess of two inches of
rainfall overflow Into the outer can,  and all measurements of liquid and melted
precipitation are made in the measuring tube with a measuring stick.

    The automatic wet/dry precipitation collector, available In several
designs, represents a specialized nonrecording instrument designed for programs
involving the chemical and/or radioactive analysis of precipitation.  The
collector is built with a sensor that detects the onset and cessation of
precipitation and automatically releases a lid to open and cover the collector.
In one design, the lid can be made to remain open during either wet or dry
periods.  Another model is made with two collectors;  the lid is made to cover
one bucket during periods of rain and snow (Figure 4.4.2).  In equipment of
this kind involving precipitation chemistry, the volume of water In proportion
to the constituents collected with the water is important, so evaporation must
be kept to a minimum  (see EPA, 1985).

       4.4.1.2 Recording Gages

Recording gages are of two basic designs based on their operating principles:
the weighing-type gage and the tipping bucket-type gage (Figure 4.4.3).  The
former, when made to NWS Specification No. 450.2201,  Is known as the Universal
gage, indicating usage for both liquid and frozen precipitation.  There are
options for the remote transmission of signals from this type of gage.  The

-------
                                                              Section No.  4.4.0
                                                              Revision No.     0
                                                              Date:    17 Sep 89
                                                              Page     2  of  7
standard Mational Weather Service Tipping Bucket Rain Gage  is  designed with a
12-inch collector funnel that directs the precipitation  to  a small  outlet
directly over two equal compartments, or buckets, which  tilt in sequence with
each 0.01 inches of rainfall.  The motion of the buckets causes a mercury
switch closure.  Normally operated on 6 V d.c., the contact closure can be
•ooltored on a visual counter and/or one of several recorders.   The
digital-type impulse can also be used with computer-compatible equipment.
Figure 4.4.1  A typical non-recording
              Rain Gage (Belfort
              Instrument Co.)
                  .
Figure 4.4.2 Automatic
vet/dry precipitation
collector.
 Bucket
 Platform
                                         Tipping
                                         Bucket
Figure 4.4.3  A Typical Weighing Rain Gage (left) and Typical  Tipping
              Bucket Rain Gage (Belfort Instrument Company).

-------
                                                             Section No.  4.4.0
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page     3  of  7
       4.4.1.3  Instrument Characteristics
       The aost accurate precipitation gage is the indicating-type gage.
However,  the recording-type gage measures the time of beginning and ending of
rainfall  and rate of fall.  The Universal weighing gage incorporates a chart
drum that is made to rotate by either an 8-day spring-wound clock or a
battery-powered clock.   Recent developments include a unit with a quartz
crystal mechanism with gear shafts for a wide range of rotation periods from
half a day to one month.

       The weighing gage is sometimes identified by the name of Its designer
(Fergusson) and comes with one of two recording mechanisms.  In the single
traverse  unit, the pen moves from the base of the drum to the top, typically a
water equivalent of 6 inches.  In a dual traverse unit, the pen moves up and
then down for a total of 12 inches of precipitation.  A variation of the
weighing  gage, a "high capacity" design with dual traverse, will collect as
much as 760 mm or 30 inches.  To minimize the oscillations incurred by strong
winds on  the balance mechanism, weighing gages are fitted with a damper
iaaersed  in sllicone fluid.  By incorporating a potentiometer in the mechanism,
the gage  is capable of providing a resistance or, as another refinement, a
voltage proportional to the amount of precipitation collected.  Linearity of
response  is usually a factory adjustment involving the use of calibrated
weights to simulate precipitation amounts.  In spite of manufacturer's
specifications, it is doubtful that the gage can resolve 0.01 inches,
especially when the bucket Is nearly empty.

       In the tipping bucket gage, the balance of the buckets and the leveling
of the bucket frame are critical.  Low voltage at the gage is imperative for
reasons of safety.  Power  is typically 6 V d.c..  The signal is provided by a
switch closure each time the bucket assembly tips (0.01 inches of rainfall per
bucket).   Rain rates are calculated from an event recorder with pens energized
sequentially to improve resolution.  The tipping bucket (a mechanical device)
takes time to tilt from one position to the next.  When the rate of fall Is
high, there is spillage and the unmeasured precipitation falls into the
reservoir.  Where there is a need for greater accuracy, the collected water is
measured manually, and excess amounts are allocated proportionately in  the
record.  The accuracy of the gage is given as 1 percent for rainfall rates of 1
in/hr or  less; 4 percent for rates of 3 in/hr; and 6 percent for rates  up to 6
in/hr.

       4.4.1.4 Accessories - Windshields and Heaters

       Accuracy of measurement for all types of gages  Is Influenced perhaps
more by exposure than by variations in design.  Windshields represent an
essential accessory to  improve the catch of precipitation, especially snow in
windy conditions.  The  improved Alter design, made of  32 free-swinging  but
separated leaves supported  1/2 inch above  the level of the gage collecting
orifice,  is an effective way to improve the catch.  In a comparison of  shielded
and unshielded 8-inch gages, it has been shown  that at a wind speed of  5 mph,
the efficiency of the unshielded gage decreases by 25  percent, and at 10 mph,
the efficiency of the gage decreases by 40 percent  (WeLss, 1961).

       In below freezing conditions when the catch  in  a gage  is snow or some
other form of solid precipitation, it  is necessary  to  remove  the

-------
                                                             Section No.  4.4.0
                                                             Revision No.     0
                                                             Date:    17 Sep 89
                                                             Page     4  of  7


collector/funnel of nonrecording  gages and the funnel in recording gages.   Some
instruments are available with built-in heater elements that are
thermostatically controlled.   An  effective heater for conditions that are not
too severe is an incandescent lamp  Installed in the housing of the gage.
Caution should be exercised,  however, as  too great a heat will result in
evaporative loss.

4.4.2  SPECIFICATIONS

       4.4.2.1 Precipitation Data Requirements

       In research studies,  especially those related to acid rain, the
instrument used most frequently is  the Automatic Precipitation Collector with
one or two collecting  buckets and a cover to prevent evaporation.  In
operational activities,  the choice  is between the weighing gage and the tipping
bucket gage.  For climatological  surveys, the choice might include one of the
above gages as well as a nonrecording type gage.  The use of a windshield is
recommended to minimize the errors  that result from windy conditions if the
application requires maximum accuracy.

       The precipitation measurement made in air quality monitoring stations is
frequently used for descriptive purposes  or for episodal analysis.  If the
effort required to achieve the level of accuracy specified by most
manufacturers of electrical recording gages is more than the application of the
data can Justify, a  tolerance of 10 percent may be adequate.

       4.4.2.2 Procurement

       In purchasing a suitable precipitation measuring system, specify the
type that fits  the data application and  include a requirement for accuracy
consistent with  that application.  A variety of gages are available
commercially.   In general, the standards  established by NWS specifications
result in the fewest problems.  For example, there are numerous 8-inch gages
available, but  those following NWS  specifications are made only of brass and
copper, are more durable, and are reported  to rupture less frequently under
extended freezing conditions than those made of galvanized steel.

       The procurement of a weighing type gage should include a tripod mounting
base as well as a set  of calibration weights.  For locations that may not be
readily accessible,  or for locations with heavy precipitation, the bucket of
the weighing gage should have an overflow tube.  Refer to Section 4.4.3.2 for
antifreeze specifications.   If the  resolution of time is not too important,
recording rain  gages of the drum type  can be obtained with monthly rather than
weekly mechanisms.   Unless the tipping  bucket gage is equipped with a heater,
it  is of no use  for  frozen precipitation.

       4.4.2.3  Acceptance Testing

       Except  for  visual inspection, nonrecording gages do not require
acceptance  testing.   The weighing gages  should be assembled and given a quick
"bench-top" calibration check using standard weights or a measured volume of
water.   In addition,  the clock mechanism supplied with the gage should be
checked  for at  least a couple of days,  preferably a week.  The tipping bucket

-------
                                                             Section No. 4.4.0
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page     5  of  7

gage should also be bench tested, primarily to be certain that the bucket
mechanism assembly is balanced and that the switch is operational.

       4.4.2.4 Calibration

       Bench calibrations should follow the recommendation of the manufacturer.
The electrical output gage or the drum recording gage measures weight, whether
total weight in the case of the "weighing gage" or increments of weight In the
case of the tipping bucket gage.  Density of water is assumed so the weight can
be expressed in units of volume or depth assuming the area of the collector
opening.   Calibrations of the measurement apparatus can be based upon the
introduction of known volumes of water.  The area of the collection surface
must be known for the volume collected to be expressed as a depth.  For
example,  an "eight inch" collector may feed a tipping bucket which tips when
7.95 cc of water has arrived.  If this volume of water is to represent 0.01" of
rainfall, the effective collection area must be 48.51 square Inches, from the
following calculations:

                  7.95 cc - 0.485 in.3  • 0.01  in.  • 48.51  in.2
If the area is a circle, the diameter should be 7.86 Inches.

                         (48.51/n)1'2 • 3.93 in. radius

For rate-sensitive systems such as the tipping bucket, the rate of simulated
precipitation should be kept less than one inch per hour.   Calibrations require
properly leveled weighing systems (gages) whether on the bench or in the field.

4.4.3     OPERATIONS

       4.4.3.1 Installation

       Refer to Section 4.0.4.4.2.4 which provides some siting guidance for
precipitation measurement.  The support,  or base,  of any gage must be firmly
anchored, preferably on a level surface so that the sides of the gage are
vertical and the collector is horizontal.   The collector can be checked with a
carpenter's level placed at two intersecting positions.   The level of the
bucket assembly on the tipping bucket gage is also critical and should also be
checked along its length and width.

       Once the weighing gage is installed, the silicone fluid should be poured
into the damping cylinder as required.   The pen of the drum recording type is
inked to less than capacity because the ink used is hygroscopic and expands
with increasing humidity, easily spilling over the chart.   The final
calibration check with standard weights or suitable substitute should be made
at this time.  To check the operation of the tipping bucket, the best approach
is to put a known quantity of water in a can with a small  hole so that the slow
flow can be timed.  It may be necessary to adjust the set  screws, which act as
limits to the travel of the tilting buckets.   The average  of a minimum of ten
tips should be used.   Adjustment is required if a 10 percent or greater error
is found or if greater accuracy is needed.

       4.4.3.2  Field Operation of a Precipitation Measurement System

       Calibration checks for weighing and tipping bucket  gages using the
techniques described above are recommended at 6-month intervals.   Nonrecording
gages,  whether alone or in a network,  should be read daily at a standard time.

-------
                                                             Section No.  4.4.0
                                                             Revision No.     0
                                                             Date:   17 Sep 89
                                                             Page     6  of  7


       Although the weighing gage  is used for  liquid and frozen precipitation,
it requires some special attention for winter  operations.  First, the funnel
must be removed when snow  is expected.  Second,  the bucket must be charged with
an antifreeze. 24 oz of ethylene glycol mixed  with 8 oz of oil.  The weight of
this mixture represents the baseline from which  precipitation amounts are to be
noted.   The bucket should  be emptied and recharged when necessary, at about 5
inches in the Universal gage,  and  at about  10  inches in the punched tape gage.
If the antifreeze mixture  is classified as  hazardous or environmentally
sensitive, care must be taken  in the disposal  process.  All operational
activities should be recorded  in the station log.

       4.4.3.3 Preventive  Maintenance

       Possible leaks  in the measuring  tube or the overflow container of the
gage are easily checked.   The  receptacles are  partially filled with water
colored with red ink and placed over a  piece of  newspaper.  This procedure is
especially applicable  to  the clear plastic  4-inch gage which  is more easily
damaged.  Repairs are  performed by soldering the 8-Inch gage  and by applying a
solvent to the plastic.

       A number of pens,  some  with greater  capacity  than others, can be used
with the Universal gage.   All  require  occasional cleaning,  including a good
soaking and wiping in  a mixture of water  and detergent.  After  inking problems,
the next source of trouble is  the  chart drive; but  these problems can sometimes
be avoided by having  the  clock drive  lubricated for  the environmental
conditions expected.   It  is a  good practice to have  spare  clocks  in stock.

       Routine visual  checks of the performance of weighing type  gages should
be made every  time there  is a  chart change.  The time  and  date  of change, and
site location  should  be documented.   Routine maintenance should include inking
the pen and winding  the  clock.  Battery-powered chart  drives  will require
periodic replacement  of  batteries  based on either experience  or manufacturer's
recommendations.   All  preventive maintenance activities  should  be noted in  the
log book.

4.4.4  PERFORMANCE AUDIT METHODS

    Audits on precipitation measuring systems need be  no more frequent  than
every  6 months.   The irregular occurrence of precipitation makes the use of  a
CTS  impractical.   The performance audit should depend  upon the  challenging  of
the gage with amounts of water known to an accuracy of at  least 1 percent of
the  total  to  be  used.   This method will provide an accuracy of  the  measurement
system but not the collection efficiency of the gage in natural precipitation.
For  tipping  bucket gages use a rate of less than one inch  per hour  and  an
amount which will cause a minimum of ten tips.

     For weighing gages,  it is more convenient to use calibration weights  to
challenge  the weighing mechanism  rather than using the gallons of water
necessary for full scale  testing.

     All  types of precipitation gages should be  measured to determine  the
effective collection area.  This  measurement  is only required once  but  the
difficulty of measuring the area  of a slightly  out-of-round collector  may
 require several samples to accurately find  the  area.

-------
                                                             Section No. 4.4.0
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page     7  of  7
4.4.5     REFERENCES
EPA.  1985:   Quality Assurance Handbook for Air Pollution Measurement Systems,
    Vol.  V. Precip. Measurement Systems, EPA-600/4-82/042a.  Office of Research
    and Development. Res. Triangle Park, NC 27711.

NCAR,  1984:  Instructor's Handbook on Meteorological Instrumentation, F. V.
    Brock,  Editor. NCAR Technical Note, NCAR/TN-237+1A.

Weiss,L -L. , 1961:  Relative catches of snow in shielded and unshielded gages
    at different wind speeds. Monthly Veather Review, Vol.  89.

WHO,  1971:   Guide to meteorological instrument and observing practices. World
    Meteorological Organization No. 8TP3, 4th edition,  Geneva, Switzerland.

-------
                                                              Section No.  4.5.0
                                                              Revision No.     0
                                                              Date:     6 Aug 89
                                                              Page:     1  of 10

        QUALITY ASSURANCE FOR RELATIVE HUMIDITY OR DEW POINT TEMPERATURE

 4.5.0   INTRODUCTION

        Humidity is  a general  term for  the water-vapor content of air.   Other.
 •ore specific,  terms for humidity include:  absolute  humidity,  relative humid-
 ity, specific humidity,  mixing ratio,  and dew point  (Huschke,  1959).   This
 section discusses the measurement of relative humidity and dew point.   Relative
 humidity  (RH) is a  dimensionless ratio of the actual  vapor pressure  of air to
 the saturation  vapor pressure at a given dry bulb temperature.   Dew  point  Is
 the temperature to  which air  must be cooled, at  constant pressure and constant
 water vapor content,  to  beosaturated with respect to  liquid water.   Frost  point
 is the  temperature  below 0  C  at which  air Is saturated with respect  to Ice.
4.5.1
TYPES OF INSTRUMENTS
       There  are many ways  to measure  the water vapor  content  of  the atmo-
 sphere.  These  can be classified in  terms of the  six physical  principles
 (Middleton'and  Spllhaus,  1953) listed  In Table 4.5.1.  Examples of  Instruments
 for each technique are provided.

                Table 4.5.1  Principles of Humidity Measurement
Principle
Reduction of temperature
by evaporation
Dimensional changes due to
absorption of moisture,
based on hygroscopic
properties of materials
Chemical or electrical
changes due to absorption
or adsorption
Formation of dew or frost
by artificial cooling
Diffusion of moisture
through porous membranes
Absorption spectra of water
vapor
'
Instrument/Method
psychrometer
hygrometers with sensors
of hair, wood, natural
and synthetic fibers
electric hygrometers
such as Dunmore Cell;
lithium, carbon, and
aluminum oxide strips;
capacitance film
cooled mirror surfaces
diffusion hygrometers
infrared and UV absorp-
tion; Lyman-alpha
radiation hygrometers
       Instruments such as diffusion hygrometers that involve the diffusion of
moisture through porous membranes are used primarily in research programs.  The
same is true of instruments that utilize the absorption spectra of water vapor,
such as infrared and ultraviolet hygrometers,  and Lyman-alpha radiation hygro-

-------
                                                             Section No. 4.5.0
                                                             Revision No.    0
                                                             Date:    6 Aug 89
                                                             Page:    2  of 10
meters.   This class of instrument requires frequent attention and represents a
major investment in procurement and maintenance costs.
       Psychrometry identifies a basic technique for
deriving both relative humidity and dew point temperature
froa a pair of thermometers—a dry bulb thermometer that
measures the ambient temperature, and a wet bulb thermo-
meter.   The reservoir of the wet bulb thermometer is
covered with a muslin wick.  When the wick is moistened
and the thermometer ventilated, the indicated temperature
is related to the amount of evaporative cooling that takes
place at the existing ambient temperature, water vapor
partial pressure, and atmospheric pressure.

       The temperature sensors in a sling psychrometer
(Figure 4.5.1) are usually mercury- or alcohol-filled
thermometers.  The same is true of portable motor-
operated psychrometers (Figure 4.3.7.3), but the psychro-
metric principle has been used with sensors made of
thermocouples, wire-wound resistance thermometers, therm-
istors, and bimetal thermometers.  Relative humidity and
dew point are easily determined by observ- ing the
difference between the dry bulb and the wet bulb—the wet
bulb depression—and then referring to psychrometric
tables, charts, or calculators.  One must be certain to
use computed values for the atmospheric pressure range of
the location where the observation is taken.
                                                                           i
                                                               ~
                                                               Figure 4.5.1
                                                                  Sling
                                                               psychrometer
       More »easurements of atmospheric water vapor have
probably been made with the sling psychrometer than by any
other manual method.   When properly used and read,  the
technique can be reasonably accurate, but it is easily misused.  The most
important errors are from radiation, changes during reading, and parallax.  The
Assaann psychrometer continuously aspirates the thermometers and protects them
                                      from radiation which allows time and
                                      accessibility for a careful reading to
                                      avoid parallax (a parallax avoiding guide
                                      to keep the eye perpendicular to the
                                      meniscus is best, see Figure 4.5.2).  For
                                      good accuracy, particularly where a
                                      variety of observers are taking measure-
                                      ments, an Assmann or equivalent type
                                      psychrometer is recommended.   One should
                                      use the psychrometric tables with dew
                                      point values for the altitude (pressure)
                                      where measurements are being made.

                                           Hygrographs, which record relative
                                      humidity, or hygrothermographs,  which
                                      record both relative humidity and temper-
                                      ature, usually incorporate human hair as
  Figure 4.5.2  Assnann psychro-      the moisture-absorbing sensor.  Other
   vzter with parallax guides.         instruments with sensors that respond to
                                      water vapor by exhibiting dimensional

-------
                                                            Section No.  4.5.0
                                                            Revision No.     0
                                                            Date:     6 Aug 89
                                                            Page:     3  of 10
changes are available.  They are made with elements  such as wood,  goldbeater's
skin (an animal membrane), and synthetic materials,  especially nylon.

       Instruments that utilize the hygroscopic  characteristics of human hair
are used most frequently, primarily because of availability.   The hygrograph
provides a direct measure of relative humidity in a  portable  instrument that is
uncomplicated and is relatively inexpensive.  There  are limitations in accuracy
below 20 percent relative humidity and above 80  percent that  may be unaccept-
able,  as well as limitations for applications at low temperatures.  Atmospheric
Environment Services of Canada has found that Pernix,  a specially treated and
flattened hair element, can be used at temperatures  below  freezing without
serious errors.  The hygrothermograph made to an NWS specification incorporates
human hair as the humidity sensor and bourdon tube (a curved  capsule filled
with alcohol) as the temperature sensor.

       Dew point hygrometers with continuous electrical outputs are in common
use for monitoring.  One dew point hygrometer was originally  developed for air
conditioning control applications under the trade name Dewcel (Hickes, 1947)
and was adopted to meteorological use (Conover,  1950).  From  the trade name,
the generic term dew cell has evolved that now  identifies  an  instrument made by
several manufacturers.  This device determines moisture based on the principle
that for every water vapor pressure there is an  equilibrium temperature at
which the saturated salt solution neither absorbs nor gives up moisture to the
surrounding atmosphere.

       The dew cell, also known by the trade name Dew
Probe, consists of bifllar wire electrodes wrapped
around a woven glass cloth sleeve that covers a  hollow
tube or bobbin.  The sleeve is impregnated with a
lithium chloride solution (Figure 4.5.3).  Low-voltage
a.c. Is supplied to the electrodes, which are not
interconnected but depend on the conductivity of the
atmospherically moistened lithium chloride for  current
flow.   The temperature sensor in the tube is usually a
resistance thermometer, but can be a thermistor.
thermocouple, bimetal thermometer, capillary system, or
any sensor calibrated for the proper temperature-to-
dew-point relationship.

       In the early 1960's, the technique of detecting
the dew point on a cooled mirror surface evolved into a
product ion-type unit.  This unit was automatically
operated and had an optical dew-sensing system that
incorporated thermoelectric cooling  (Francisco  and
Seaubien. 1963).  Four manufacturers now produce a
Meteorological type, thermoelectric, cooled-mirror dew
point instrument (Mazzarella, 1977).  Three of  these
instruments cover the range of -50 to +50 C.  Linear
thermistors are used to measure the mirror temperature
in three of the units; a platinum wire sensor is used
in the other.  All are designed with simultaneous
linear output signals for T
and T (ambient temperature).
(dew point temperature)

Two of the manufacturers
Figure 4.5.3  A
 typical Devcel
 sensor housing
and transmitter

-------
                                                             Section No. 4.5.0
                                                             Revision No.    0
                                                             Date:    6 Aug 89
                                                             Page:    4  of 10


•afce claim  to NBS-traceability with  stated dew point accuracies ranging from
±0.2° to ±0.4°C and  ambient temperature accuracies  ranging from ±0.1° to
±0.5°C.  All incorporate some form of  standardization  that involves clearing
the mirror  by heating,  either automatically or manually.  Although complex in
design and  operation,  this type of cooled-mirror  hygrometer  is considered to be
a functional standard.

       In recent  years,  two other sensors for humidity have  been used on tower
installations for atmospheric pollution studies.  One  involves the use of an
organic seed, cut and coupled to a strain gage.   In principle, absorption of
•oisture in the seed results in distortion, which is converted to an electrical
signal by the strain gage assembly.  Reports on performance  are mixed.
Certainly the applications are limited, and the approach does not represent a
technological advance.   By contrast, the thin film  capacitor, designed primari-
ly for radiosonde applications, incorporates advanced  technology  (Suntola and
Ant son, 1973).  Reports of users In  the past have been mixed, with a common
complaint of poor performance in polluted atmospheres.  Modern capacitor-type
sensors have achieved a better performance through  improved  design and user
education.

       4.5.1.1    Sensor Characteristics

       Although  the  psychrometer is  considered the  most practical and widely
used instrument  for  measuring humidity, two major problems are associated with
wet and dry bulb  psychrometry involving the accuracy of the  thermometers and
the cumulative errors related to operating technique  (Quinn,  1968).  An  accur-
acy of ±1 percent at 23°C and 50 percent RH requires thermometers with  relative
accuracy of ±0.1°C.   The commonly used 0.5 C division  thermometers introduce an
uncertainty of  ±5 percent RH at this condition.   This  assumes  that the  readings
were taken  at  the maximum wet bulb depression, a  difficult task with a  sling
psychrometer.

       It has  long been recognized  that  there are some limitations in using the
dew cell  instrument  (Acheson, 1963).  The  lowest  relative humidity it can meas-
ure at a given temperature is the  ratio of  the vapor pressure  of  a saturated
solution of L1C1  to that of pure water.  This  is  calculated  to be  11.8  percent
RH.  A second  limitation is that at  -65.6°, -20.5°, +19.0°,  and +94.o"c, L1C1
in equilibrium with its saturated  solution undergoes a phase change.  Errors in
dew point measurements occur at -69°,  -39°, -12°, and  +40°C.   This problem is
inherent  in the use of LiCl and cannot be eliminated.   It  is estimated  that the
accuracy of the LiCl saturated salt  technique  is  1.5°C over  the range of -30°
to 30°C.

       The  optical chilled (cooled)  mirror  technique of measuring dew point is
a fundamental  measurement.  No calibration  is  required for the fundamental dew
generating  process.   The measurement however  is  the temperature of the  surface
at which  the dew forms and as with any electrical temperature  measurement
system, calibration is required.  The process  of  periodically heating  the
mirror to  a temperature above  the  dew (or  frost)  point is  followed by zeroing
the  optical system to correct for  the dry  mirror  reflectance changing due  to
contamination.    In the better  instruments,  automatic  zeroing is programmable in
terms  of  frequency and  length of time.  It  can also be accomplished manually.

-------
                                                             Section No.  4.5.0
                                                             Revision No.     0
                                                             Date:     6 Aug 89
                                                             Page:     5  of 10
       4.5.1.2   Sensor Housings and Shields

       Psychrometers of all types should be acclimated to the environmental
conditions in which the measurements are to be made.   In most cases,  psychro-
•eters should be stored in a standard instrument shelter so that the mass of
the thermometers,  especially the mass of the housing,  adjusts to the temper-
ature of the air.   Psychrometers with a stored water supply,  such as those on a
tower, must be shielded from solar radiation.

       For meteorological applications, the dew cell element should be enclosed
in a weatherhood to protect it from precipitation,  wind, and radiation effects.
This type of element functions best in still air.   Some aspirated radiation
shields are designed, in keeping with these specifications to house both a
temperature sensor, which requires ventilation,  and a dew cell,  which requires
only the smallest amount of air flow (Figure 4.5.4).   The miniaturization of
the dew cell has created some problems related to excessive air flow and solar
radiation that remain only partially solved.

       All manufacturers of optical cooled-mirror dew point and temperature
aonitoring equipment provide housings for the sensors, which include forced
ventilation and shielding from solar radiation.

       4.5.1.3   Data Requirements

       Electrical hygrometers for monitoring applications have time constants
generally longer than air temperature systems.   The usual data of interest are
hourly average values.  Data should be reported in terms of the condition
                                       measured, dew point temperature,
                                       relative humidity or wet-bulb and dry-
                                       bulb temperature.  Programs may be used
                                       which convert among these if all the
                                       relevant variables are known.   The
                                       station elevation may be used to
                                       estimate a nominal pressure if a
                                       measurement is not available.   The temp-
                                       erature needed to convert a relative
                                       humidity measurement to dew point temp-
                                       erature is that temperature at the
                                       relative humidity sensor surface.  This
                                       may not be the same temperature as that
                                       measured at some other location.  On the
                                       other hand,  the dew point temperature Is
                                       a fundamental measure of the amount of
                                       water vapor in the air and is indepen-
   Figure 4.5.4  A pair of tower-      dent of air temperature.   Relative
   mounted Gill aspirated radiation    humidity calculations can therefore be
   shields for housing temperature     made given the dew point temperature and
   and dew point sensors (Young).       any temperature measurement point in the
same general air mass.  Empirical formulae for the estimation of relative
humidity as a function of dew point temperature and air temperature,  relative
humidity as a function of wet-, dry-bulb temperature and pressure, and dew
point temperature as a function of relative humidity and temperature are shown
below.

-------
                                                             Section No.  4.5.0
                                                             Revision No.     0
                                                             Date:     6 Aug 89
                                                             Page:    6  of 10
       To calculate relative humidity  (RH = 100 r, %) from air temperature  (T,
°C)  and dew point temperature  (T  , °C), do the following:
                    exp
        (• t
                                                            (1)
                     where a =   17.27
                           b = 237.3
       To calculate the dew point temperature (T ,  °C) from air temperature (T,

*C) and relative humidity  (RH =  100 r, %) use
                               b  In r
                                          b + Tl
                             a -  In r +
                           aT
                          b + T|
                                                            (2)
       To calculate relative humidity  (RH = 100 r, '/•) from air temperature
(T. °C), wet-bulb temperature (T ,  °C), and atmospheric pressure (P, mb)
through the vapor pressure  (e.  mb)  and  the saturation vapor pressure (e ,  mb),
do the following:
                     e
                     e
                                       + BV(T - V
                                           (3)
                              eoexp
      where A

            B
6.6  x 10

1.15 x 10
                    exp
                          -4
-3
                            aT
                           b + T
       To estimate  wet-bulb temperature  (T  ,  C) from air temperature (T, °C),

dew point temperature  (T ,  °C).  relative humidity  (r, ratio) and atmospheric

pressure (P,   mb),  do  the following:
Tw£T
P H
P -
hTo
^ T
19 +
19 +
130r
130r
-28r2
-28r2

                                                            (4)
The summarization of  these relationships was suggested by A. L. Morris from
material found  in Z.  Ceophysik,  6,  297,  1930,  the Smithsonian Meteorological
Tables, Sixth Revised Edition,  the  Glossary of Meteorology and has been
augmented by his own  derivation of  expression  (4).

-------
                                                             Section No. 4.5.0
                                                             Revision No.    0
                                                             Date:    6 Aug 89
                                                             Page:    7  of 10


       Psychrometers are convenient devices for making spot checks of the per-
formance of other devices, especially those that are permanently  installed,
providing the checking is done under reasonably steady overcast conditions.
The psychrometric technique built into tower Installations presents servicing
problems, especially at temperature extremes.  High temperatures  cause rapid
evaporation, and low temperatures cause freezing.

       Both the dew cell and the cooled-mirror type instruments have applica-
tions on lO-meter or taller tower installations for pollution studies,
providing the sensors are housed in the recommended shields with  little, if
any, aspiration for the dew cell and the recommended rate of aspiration for the
cooled-mirror design is selected.

4.5.2  SPECIFICATIONS

       4.5.2.1   Procurement

       The selection of a humidity instrument is guided by the application to
which the data will be put.  The PSD (Prevention of Significant Deterioration)
guideline (EPA. 1987) provides the following:  "...If the permit  granting auth-
ority determines that a significant potential exists for fog formation,  icing,
etc.. due to effluents from the proposed facility, error in the selected
measurement technique should not exceed an equivalent dewpolnt temperature
error of 0.5°C.  Otherwise, errors in equivalent dew-point temperature  should
not exceed 1.5°C over a dewpolnt range of -30°C to +30°C."  This  latter  toler-
ance allows for the use of lithium chloride dew cells.

       Sling psychrometers and aspirated psychrometers with thermometers
snorter than 10 inches do not have sufficient resolution for the  accuracies
required for checking other instruments.—Equally important, the  thermometers
should have etched stems; i.e., the scale markings should be etched on  the
glass.  Reliable thermometers are factory calibrated at a minimum of  two temp-
eratures, and usually at three.  Thermometers calibrated with NBS-traceable
standards are preferred.

       When patents expired on the original Dewcell. a number of  similar units
appeared on the market.  In light of problems which have existed  in the  past,
it is prudent to specify accuracy of the humidity system when it  is operating
as a systea in the atmosphere.  Problems with ventilation rates will  be  quickly
exposed by this requirement.  It is not recommended to purchase components to
patch together in a system.  Corrosion in polluted atmospheres can be avoided
by selecting optional 24-carat gold windings, provided cost is not prohibitive.
If dew point alone is to be measured, the standard weatherhood is a proper
choice.  If both temperature and dew point are to be measured, it may be
advantageous to purchase a standard shield that provides a housing for  the dew
cell and a separate aspirated compartment for the temperature probe.

       Optical cooled-mirror dew point systems are now commercially available
from several manufacturers, all of which incorporate either linear thermistors
or platinua resistance  temperature devices.

-------
                                                             Section No. 4.5.0
                                                             Revision No.    0
                                                             Date:    6 Aug 89
                                                             Page:    8  of 10
       4.5.2.2   Acceptance Testing
       Test  at  least  the ambient atmosphere at one  point in normal wind and
radiation.

       4.5.2.3    Calibration

       The procedure  for calibrating the thermometers  in a psychrometer is
essentially  the same  as any thermometer calibration (See Section 4.3.5).

       Both  the dew cell and the cooled-mirror hygrometer can be checked for
approximate  calibration accuracy with a motor-operated psychrometer.  Their
performance  should be verified under stable conditions at night or under cloudy
conditions during the day.   Several readings taken  at  the intake of the aspir-
ator or  shield  are recommended.  Bench calibrations of these more sophisticated
units must be made by the manufacturer.  The electronics portion of some
instruments  may be calibrated by substitution of known resistances in place of
the temperature sensor.  This procedure, if appropriate, is described in the
   lufacturer's  operating manual for the instrument.
4.5.3  OPERATIONS

       4.5.3.1   Installation

       Dew point measuring equipment on a tower should be installed with  the
same considerations given to temperature sensors.   Reference has already  been
•ade to  the weatherhood as a shield for the dew cell and to an aspirated  shield
for the  cooled-mirror instrument.  At some installations, -success has been
reported in mounting these housings so that they are close to the tower frame-
work on  the north-facing side.  This minimizes the effects of direct solar
radiation and provides a rigid support, especially for the coolea-mirror
sensor,  which requires a stable mounting surface.   Another consideration  in
Mounting these devices inboard involves servicing.  Inboard mounting makes
recharging the dew cell with lithium chloride and cleaning the reflective
surface  of the cooled- mirror hygrometer much easier.

       4.5.3.2   Field Operation and Preventive Maintenance

       Field calibration checks should be made at least monthly on dew cell
type units.  The use of gold wire windings around the LiCl cylinder minimizes
corrosion problems in polluted atmospheres.  Periodic removal and washing of
old  lithium chloride, followed by recharging with a fresh solution, improves
data reliability.

       Once a mercury or alcohol liquid-in-glass thermometer is calibrated,
there  is no need for recalibratlon, unless it is to be used for reference or as
a transfer standard.  Errors  in wet bulb temperatures are most frequently the
result of an improperly installed or dirty muslin wick, the repeated use  of  tap
water  Instead of distilled water, or human error in reading.  Ulcking material
used on psychrometers must be washed to remove traces of sizing and finger-
prints.   Once cleaned, the material is tied at the top of the thermometer bulb
and  a  loop of thread placed around the bottom so the thermometer bulb  is
 tightly covered.  To prevent  solid materials from collecting on the cloth and
preventing proper evaporation, the wick should be wet with distilled water.  Of

-------
                                                             Section No.  4.5.0
                                                             Revision No.    0
                                                             Date:     6 Aug 89
                                                             Page:     9  of 10

course,  slinging or motor aspiration should be done In the shade, away from
reflected or scattered radiation,  at a ventilation rate of about 3 to S m/s.
Many technique-related errors are minimized by using an Assmann-type, motor-
operated psychrometer. providing the instrument is allowed to assume near
ambient conditions prior to use.

       The cooled-mirror instruments require no calibration except for the
•inor temperature sensor.  Depending on environmental conditions, the mirror is
easily cleaned with a Q-Tlp dipped in the recommended cleaning fluid, usually a
liquid with an alcohol base.  While the accuracy of a psychrometer is inferior
to that of the optical chilled mirror system, an occasional check at the intake
to the sensor shield is recommended under the provisions specified earlier.

       All operational and preventive maintenance activities should be logged.
Data retrieval will be dependent upon program objectives.

4.5.4  PERFORMANCE AUDIT METHODS

       Instrument audit procedures for hygrometry systems follow calibration
procedures.  A systems audit should be performed near the beginning of a field
Measurement program.

       The performance audit of a humidity measuring system should be based
upon a comparison with a collocated transfer standard (CTS).  Parts of the
system can be tested by conventional electronic tests, but this avoids so much
of the measurement process that it should only be used to augment the total
system test.  The CTS may be any qualified Instrument.  The most accurate type
Is the cooled-mirror dew point instrument.  The Assmann-type psychrometer with
calibrated thermometers traceable to NBS is acceptable for most data applica-
tions.  It Is also most convenient since it does not require commercial power
and can be carried to elevated levels on a tower.

       Given the qualifier that humidity is a very difficult measurement to
aake, a rule of thumb for judging the accuracy of a humidity monitoring system
with an As smarm as the CTS is as follows:  when the CTS and the challenged
system agree In dew point temperature to within 1 C, the challenged system  is
assumed to be within 0.5°C of the true value;  This arbitrarily assigns an
uncertainty in dew point temperature of +0.5°C for the Assmann which is true
for most of the range.

       Auditing is best backed by authoritative standards.  ASTM, 1982, 1983,
1984 and 1985 may be of selective value.

-------
                                                             Section No. 4.5.0
                                                             Revision No.    0
                                                             Date:    6 Aug 89
                                                             Page:   10  of 10
4.5.5  REFERENCES
Acheson, D. T. ,  1963:  Some limitations  and errors  inherent  in  the use of
    dew-cell for measurement of  atmospheric dew  points. Monthly Veather Review.
    ASTM, 1982:  Standard  Definitions  of  Terms Relating  to Humidity
    Measurements,  D4023-82a, American Society for Testing and   Materials,
    Philadelphia.  PA.

ASTM. 1983:  Standard  Method of  Measuring Humidity  with CooLed-Surface
    Condensation (Dew  Point) Hygrometer,  D4230-83.  American  Society  for
    Testing and Materials, Philadelphia. PA.

ASTM. 1984:  Standard  Test Method for Measuring  Humidity with a Psychrometer
    (the Measurement of Vet- and Dry-Bulb Temperatures), E337-84, American
    Society for Testing and  Materials,   Philadelphia,  PA.

ASTM. 1985:  Standard  Practice for Maintaining Constant Relative Humidity by
    Means of Aqueous Solutions,  E104-85,  American Society  for  Testing and
    Materials,  Philadelphia. PA.

Berry.  F. A. ,  Jr.. E.  Bollay and N.  R. Beers,  1945:  Handbook of Meteorology.
    McGraw-Hill Book Company. Inc.

Conover, J. H., 1950:   Tests and adaptation of  the  Foxboro dew-point recorder
    for weather observatory use. Bulletin of American  Meteorological
    Society.31(1).13-22.

EPA,  1987:  Ambient Monitoring Guidelines for Prevention of  Significant
    Deterioration (PSD).  EPA-450/4-87-007. Office of Air  Quality  Planning and
    Standards,  Research Triangle Park, NC.

Francisco  and Beaubien,  1965:  An automatic dew point hygrometer with
     thermoelectric cooling. Humidity and Moisture,  edited by A.Wexler,
    Relnhold  Publishing Company.

Hicks.  W.  F. .  1947:  Humidity measurement by a new system.  Refrigerating
    Engineering. American  Society of Refrigerating Engineering.

Huschke. R. ,ed..  1959:  The Glossary of Meteorology. American  Meteorological
     Society.  Boston,  MA.

Mazzarella.  D. A.  . 1972:    Meteorological instruments:  their selection and  use
     in  air pollution studies. Proceedings of the Meeting on  Education and
     Training in Meteorological  Aspects of Atmospheric  Pollution and Related
     Environmental Problems. World  Meteorological Organization, No.  493.

Middleton,  W.  E.  K. , and  A. F.  Spilhaus,  1953:   Meteorological Instruments,
     University  of Toronto  Press.

Quinn.  F.  C. ,  1963:  Humidity-the neglected parameter. Testing  Engineering.
     The Mattingly Publishing  Company, Inc.

 Suntola aud Antson, 1973:   A  thin film humidity sensor. Scientific
     Discussions,  CIMO VI,  World Meteorological  Organization.

-------
                                                             Section No.  4.6.0
                                                             Revision No.     0
                                                             Date:    17 Sep 89
                                                             Page:      1   of 8

               QUALITY ASSURANCE FOR SOLAR RADIATION MEASUREMENTS

4.6.0  INTRODUCTION

    Solar energy Is the driving force of  large-scale atmospheric motion,
indeed, of the general circulation of the atmosphere.  Although air pollution
investigators normally consider the  measurement of  solar radiation secondary to
wind and temperature measurements, solar  radiation  Is directly related to
atmospheric stability.  It is measured  as total incoming global radiation, as
outgoing reflected and terrestrial radiation and  as net total  radiation.

    Quantitatively, solar radiation  is  described  in units of energy flux.
either W/m  or cal/cm «min.   When measured in specific, narrow wavelength
bands, solar radiation may be used to evaluate such air pollution indicators as
turbidity, amount of preclpltable water,  and rates  of photochemical reactions.
However, this manual will cover only broadband measurements and sunshine.

    The generic term, radiometer, refers  to any Instrument that measures
radiation, regardless of wavelength.  Shortwave radiation has  wavelengths less
than 4 micrometers (fim) and  is subdivided as follows:
                      Ultraviolet (UV)
                      Visible
                      Near-infrared
                                        0.20 tun to 0.38 fin
                                         0.38  fim to  0.75 fim
                                         0.75  fim to  4.00 fim
    Longwave radiation has a wavelength as follows:

                      Infrared (IR)     4    fim to  100 fim

and comes from the Earth and its atmosphere.   The  instruments most commonly
used for environmental monitoring  are  discussed below.

4.6.1  TYPES OF INSTRUMENTS
       4.6.1.1 Pyranometers

       Pyranometers are Instruments  that measure  the  solar  radiation received
from the hemispherical part of
the atmosphere It sees,
including the total sun and
                                            Sensing
sky shortwave radiation on a
horizontal surface (Figure
4.6.1).   Most pyranometers
incorporate a thermopile as
sensor.   Some use a silicon
photovoltaic cell as a sensor.
The precision spectral
pyranometer (PSP) is made by
Eppley Laboratories and has
two hemispherical domes
designed to measure sun and
sky radiation on a horizontal
surface In defined
wavelengths.  This is achieved
by substituting one of several
                                              Element
                                       Guard Disc
                                           \
                 Precision Ground &
                 Polished Glass Dome
                 (Frequently Double)
                                                                         Leveling
                                                                          Screw
                                 Figure 4.6.1
Features of a. typical
pyranometer (Carter, et al.,1977)

-------
                                                             Section No. 4.6.0
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:     2  of 8


colored Schott glass filter doues for the clear glass outer dome.  The smaller
done suppresses convection, so this type is better sited if tilted from the
horizontal.

       4.6.1.2 Bimetallic Recording Pyranometers

       Bimetallic recording pyranometers, also known as actin- ometers,
predesigned by Robitzsch of Germany,  these mechanical sensors consist of two
or three bimetallic strips, alternately painted black and white,  that
respectively absorb and reflect solar radiation.  The resulting differential
heating produces a deformation that is transmitted mechanically through levers
and a pen arm to a clock-wound drum recorder.  Although of limited accuracy,
these instruments are useful for locations with no commercial power.

       4.6.1.3 Net Radiometers

       Net radiometers or net pyrradiometers are designed to measure the
difference between downward and upward total radiation, including the total
incoming shortwave and longwave radiation and the
total outgoing shortwave and longwave radiation.
There are two basic types of net radiometers.  The
ventilated plate type, often referred to by the
Dane of the designers (Gier and Dunkle), is more
popular in research applications than the type
with hemispherical polyethylene domes originally
designed by Funk.  Both incorporate thermopiles
with blackened surfaces.  Because net radiometers
produce a signal with a positive sign when the
incoming radiation exceeds the outgoing, the
recording equipment must be designed with an
offset zero.

       4.6.1.4 Sunshine Recorders

       Sunshine recorders are designed to provide
information on the hourly or daily duration of
sunshine.  Only one commercially available,
off-the-shelf type of sunshine recorder  is now
available.  This is the Campbell-Stokes  design
(Figure 4.6.2), designated as the interim
reference sunshine recorder "IRSR" by the World
Meteorological Organization.  The device consists
of a glass sphere 10 cm in diameter mounted in a
spherical bowl.  The sun's rays are focused on a
card that absorbs radiation and changes  color in
the presence of sunlight.  The recorder  is used
infrequently in the United States but extensive-
ly abroad, primarily for the collection  of
climatological data.  The National Weather Service
routinely uses a Sunshine Switch, which             Figure 4.6.2  A Campbell-
incorporates one shaded photocell and one exposed   Stokes Sunshine Recorder
photocell.                                          (U.S. Army,  1975)

-------
                                                            Section No. 4.6.0
                                                            Revision No.    0
                                                            Date:   17 Sep 89
                                                            Page:     3  of 8
   Pyrhetiometer
Declination Adjustment

       24 hr. Dial
       4.6.1.5 Instrument Characteristics

       Only the characteristics of  pyranometers and net radiometers,  the two
types of instruments used most frequently  in pollution-related programs,  will
be discussed in this section.   The  pyranometer is not  to be  confused  with the
pyrhellometer, "an instrument  for measuring the intensity of direct solar
                                        radiation at  normal incidence"  (WHO.
                                        1971).  The pyrheliometer is mounted
                                        in a solar tracker,  or equatorial
                                        mount, automatically pointing to the
                                        sun as It traverses from east to west
                                         (Figure 4.6.3).  By contrast, the
                                        pyranometer is mounted facing toward
                                        the zenith.   Ideally, the response of
                                        the thermopile sensor in the
                                        pyranometer is proportional  to the
                                        cosine of the angle of the solar beam
                                        and is constant at  all azimuth angles.
                                        This characteristic is known as the
                                        Lambert Cosine Response, an important
                                        characteristic of pyranometers.  For
                                         the majority  of applications related
                                         to atmospheric pollution, Class 2 and
                                        Class 3 are satisfactory (see Table
Figure 4.6.3  Features of a             4.6.1).
typical pyrheliooeter and track-
ing mount (Carter, et al., 1977)                Most  net  radiometers now
                                        available commercially are made with a
small disc-shaped thermopile covered by polyethylene hemispheres.  In most
units the material used for shielding the  element  from the wind and weather  is
very thin and is transparent to wavelengths of 0.3  to  60 pm.  Until recently,
the internal ventilation and positive pressure required  to maintain the
hemispheres of net radiometers in their proper shape was considered critical;
however, new designs have eliminated this  problem.  The plate-type net
radiometer, most often the modified Gler and Dunkle design sold commercially in
the United States, is occasionally used In routine air pollution
investigations.  The thermopile heat flow transducer  is blackened with a
material that is easily cleaned with water or naphtha. Because the thermopile
is uncovered for total spectrum response,  a built-in  blower, available for
operation on 115 V 50/60 Hz or 12 V d.c.,  draws  air across  the element at a
constant rate eliminating the effects of varying natural winds.  The device  is
temperature-compensated and typically has  a sensitivity of 2.2 fiV per W/m ,  a
response time of 10 seconds, and a "relative" accuracy of  two percent in
calibration.  When supplied with a reflective shield  on its  lower surface, this
plate type net radiometer of the Gier and Dunkle design becomes a total
hemispherical radiometer or unshielded pyranometer.

       4.6.1.6  Recorders and Integrators for Pyranometers and
                Net Radiometers

       The relatively high impedance and low signal  of thermopile sensors,
excluding silicon photovoltaic cells, limits their use with both Indicating
meters and recording meters.  Electronic strip  chart  millivolt potentiometric
recorders incorporating variable-range rheostats are  preferred.  The

-------
                                                             Section No. 4.6.0
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:     4  of 8
variable-range rheostat permits  the exact matching of the recorder scale to
interchangeable sensors so  that  deflections of the meter represent engineering
units, i.e.. W/m2, cal/cm *min.  etc.  The alternative is a standard millivolt-
meter potentlometric recorder where the data, In millivolts, must be translated
to units of energy, corresponding to full-scale values of 1370 U/m  or 1.96
cal/cm *min.  It may also be necessary, especially if the signal is to be used
as an input to a computer,  to combine preampllfication with scaling.

4.6.2  SPECIFICATIONS

       4.6.2.1   Procurement

       In purchasing a solar radiation measurement system, follow the practice
of Batching the data requirements to the instrument selection, specifying the
performance required on the purchase order  (complete with test method to verify
performance) and testing the performance in receiving.  See Section 4.1.4.5 for
convents on traceabillty protocol.  Many types of radiation instruments have
been developed, especially  in recent years, because of an increasing interest
in environmental considerations  (Gates, 1962). meteorological research
(Monteith. 1972), and solar energy  (Carter, et al.. 1977).  Except for special
studies, the requirements for relating radiation to stability can be satisfied
by purchasing sensors of Class 2 or Class 3 as identified by the WHO (see Table
4.6.1).

       Table 4.6.1 Classification of Pyranometers According to
                        Physical Response Characteristics
Sens.
(mW/cm2)
1st Class ±0.1
2nd Class ±0.5
3rd Class ±1.0
Temp.
( '/. )
± 1
± 2
± 5
Lin.
( */. )
± 1
± 2
± 3
Max Time
Constant
(sec. )
25
60
240
Cosine
Response
( '/. )
± 3.0
± 5.7
± 10
       Class 2  sensors offer the advantage  of providing data comparable to that
collected at National  Weather Service stations and at key locations of
Department of Energy (DOE).  The sensors to  be specified should be  commercially
available, field proven by the manufacturer for several years, and have the
technical requirements established by WMO standards.  Several American Society
for Testing and Materials (ASTM) standards  are available (ASTM.  1984).  When
purchasing a recorder  or integrator,  one must match the calibration factor or
sensitivity of  the sensor to the readout equipment.  It must be  recognized that
the signals from net radiometers,  in contrast to pyranoraeters, require
zero-offset capability to accommodate both  negative and positive voltage
outputs.

       4.6.2.2  Acceptance Testing

       Physical inspection of the relatively fragile pyranometers  or  net
radiometers immediately after delivery of the instrument is  important.  One
must be sure  that the calibration data have been received and  that these data
correspond to  the serial number of the instrument.  Storage  of this  information

-------
                                                             Section No. 4.6.0
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:     5  of 8


will prove helpful when the time comes to have the calibration of the
instrument checked, or to replace the sensor or readout device.  Few
organizations are equipped or staffed to bench-test a radiometer to verify
calibration, but a quick determination can be made indoors as to whether the
sensor and recorder or integrator system is operating by exposing the sensor to
the light of a tungsten lamp.  It may be necessary to place the instrument
fairly close to the lamp.  Covering the sensor for several hours will ensure
that the system is not "dark counting."

       4.6.2.3  Calibration

       The user of a pyranometer or net radiometer is normally not equipped to
calibrate the sensor.  The best the user can do is to perform field calibration
checks on two cloudless sky days.  These checks involve a side-by-slde
comparison of the sensor to a sensor of similar design, the calibration of
which can be traced to a transfer standard.  Since 1975 all measurements have
been made in accordance with the Absplute Radiation Scale or equivalently the
World Radiometrlc Reference established at the International Pyr he Home trie
Comparison IV at Davos, Switzerland (NCAR. 1984. pp.  4-103).  If a side-by-slde
comparison is not possible, the device should be returned to the manufacturer
or to a laboratory with the facilities to check the calibration.  The frequency
of making comparative readings or having factory calibrations will depend on
environmental conditions.  Any indication of discoloration or peeling of a
blackened surface or of scratches on the hemispheres of a pyranometer warrants
recalibration and/or service.

          Net radiometers are more delicate and require more frequent attention
than pyranometers.  Pyranometers of high quality In a clean atmosphere may
require recalibration annually; net radiometers should be recalibrated at least
yearly.  Calibrating the recorder or integrator is an easy task.  The standard
nethod involves the use of a precision potentiometer to impress known voltages
into the circuit.  The linearity of the readout instrument may be checked by
introducing a series of voltages covering the full scale, checking first
up-scale and then down-scale.  Adjustments should be made as necessary.  In the
absence of a precision potentiometer, it may be possible to introduce a
calibrated millivolt source that covers one or two points.  Integrators can be
checked the same way. except that the input value must also be timed.

4.6.3  OPERATIONS

       4.6.3.1 Installation

       The site selected for an upward-looking pyranometer should be free from
any obstruction above the plane of the sensor and should be readily accessible
for cleaning and maintenance.  It should be located so that shadows will not be
cast on the device, and away from light-colored walls or other objects likely
to reflect sunlight.  A flat roof is usually a good choice; but If such a site
is not possible, a rigid stand with a horizontal surface some distance from
buildings or other obstructions should be used.  A site survey of the angular
elevation above the plane of the radiometer surface should be made through 360
degrees (The Eppley Laboratory, Inc.).

       The same procedures and precautions should be followed for nel
radiometers that are both upward- and downward-look ing.  However, the

-------
                                                             Section No. 4.6.0
                                                             Revision No.    0
                                                             Date:   17 Sep 89
                                                             Page:     6  of 8
instrument must be supported on an arm extending from a vertical support about
1 • above the ground.  Except for net radiometers with heavy-duty domes, which
are Installed with a deslccant tube in series with the sensor chamber, most
other hemispherical net radiometers require the positive pressure of a
gas—usually nitrogen—to both maintain the shape of the polyethylene domes and
purge the area surrounding the thermopile.  In on increasingly popular design,
there is a requirement for internal purging with nitrogen and external
ventilation with compressed dry air through holes on the frame.  The compressed
air supply minimizes fogging and condensation.

       Precautions must be taken
to avoid subjecting radiometers
to mechanical shock during
installation.  They should be
installed securely and leveled
using the circular spirit level
attached to the instrument.  Net
radiometers are difficult to
•ount and to maintain free of
vibration.  Pyranometers of the
Holl-Gorzynski design, used
extensively by Atmospheric
Environmental Sciences (AES) of
Canada,  are oriented so that the
emerging leads face north (Figure
4.6.4).   This minimizes solar
beat on the electrical
connections of an instrument that
is not temperature compensated.
The thermopiles of these
instruments should be oriented so
that the long side of the
         «  10  in.
Figure 4.6.4  A Moll-Gorczynskia
   Solarimeter (U.S.  Army, 1975)
thermopile points east and west (Latimer, 1972).  The cable used to connect the
pyranometer to the readout device, recorder, or integrator should be between
16 and 20 gauge and made of shielded, waterproofed 2-conductor copper wire.
The sensor, shield, and readout device should be connected to a common ground.
Potentlometric millivolt recorders are to be used with most high-impedance,
low-signal radiometers.  Cable lengths of 300 m or more are practical.
Galvanometric recorders can be used with silicon cell radiometers.   Soldered,
copper-to-copper junctions between instrument connectors and/or cables are
essential.  Pyranographs or actinographs should be installed on a level surface
iuune to shadows.   These instruments should be placed in such a way that the
sensitive bimetallic strips lie within 2 degrees of true east and west with the
glass inspection window facing north (in the northern hemisphere).

       4.6.3.2 Field Operation of a Solar Radiation System

       As part of the quality assurance program, a field calibration check
should be performed at least once every 6 months according to the procedures
outlined in Section 4.6.2.3.  Solar radiation instruments require almost daily
attention.  The data should be Inspected for a reasonable diurnal pattern and
the absence of dark counting.   Where strip chart or digital printers are used,
daily time checks are desirable.   Data retrieval will depend upon program

-------
                                                             Section No.  4.6.0
                                                             Revision No.     0
                                                             Date:    17 Sep 89
                                                             Page:      7  of 8


objectives; but even for ciimatological programs,  data should be collected
monthly.   All operational activities during a site visit should be logged.

       4.6.3.3 Preventive Maintenance

       All types of radiometers require frequent cleaning to remove any
material deposited on the surface that will intercept the radiation.  In most
cases,  this is a dally operation.  The outer hemisphere should be wiped clean
and dry with a lint-free soft cloth, using alcohol..   Any scratching of the
surface will alter the transmission properties of the glass, so cleaning must
be done with care.  If frozen snow, glazed Ice. hoarfrost or rime ice is
present,  an attempt should be made to remove the deposit carefully with warmed
cloths.

       Should the internal surface of a pyranometer's outer hemisphere become
coated with moisture, it can be cleaned by carefully removing the outer
hemisphere on a dry day and allowing the air to evaporate the moisture, then
checking the dessleant.  If removal of a hemisphere exposes the thermopile
element,  extreme care should be taken because it is fragile and easily damaged.
About once each month, the desiccant installed in most pyranometers should be
Inspected.  Whenever the silica gel drying agent is pink or white instead of
blue, it should be replaced or rejuvenated by drying it out on a pan in 135 C
oven.  The level should be checked after each servicing of the radiometer, or
at least monthly.  Significant errors can result from misalignment.

       Net radiometers require more frequent maintenance attention  than
pyranometers.  It is necessary to replace the polyethylene domes as often as
twice a year or more before the domes become discolored, distorted, or cracked.
More frequent replacement  is necessary in polluted environments due to
accelerated degradation of plastic hemispheres when exposed to pollutants.  A
daily maintenance schedule is essential to check on the proper flow of gas in
instruments that are inflated and purged with nitrogen.  All PM activities
should be  recorded in a log.

4.6.4  PERFORMANCE AUDIT METHODS
    A performance audit on a solar radiation system is only practical with a
CTS.  The  CTS must have the spectral response and exposure equivalent to  the
instrument being audited.  One diurnal cycle will establish an estimate of
accuracy sufficient for most air quality monitoring applications.   The method
of reporting the data from the monitoring instrument  (daily integrated value.
hourly integrated value, average intensity per hour. etc.) must be  used in
reducing the data from the CTS to provide a meaningful comparison.  An audit
frequency  of at  least six months is recommended.

4.6.5  REFERENCES

ASTM.  1984:  Calibration of secondary reference pyrheliometers and
    pyrheliometers for field use, E816. American Society for Testing and
    Materials, Philadelphia. PA.

Carter, E. A. et al.,  1977:  Catalog of solar radiation measuring equipment.
    ERDA/ORO/5361-1, U.S.  Energy and Development Administration.

-------
                                                              Section No. 4.6.0
                                                              Revision No.    0
                                                              Date:   17 Sep 89
                                                              Page:     8  of 8


Gates. D. M. .  1962:   Energy Exchange in the Biosphere.  Harper and Row.

La timer, J.  R..  1972:   Radiation measurement. Technical Manual Series No. 2,
    International  Field Year for the Great Lakes. Canadian National Commission
    for the  Hydrological Decade.

Monteith, J.  L..  1972:   Survey of Instruments for micrometeorology.
    International  Biological Programs Handbook Ho. 22. Blackwell  Scientific
    Publications,  Osney Mead, Oxford, England.

NCAR. 1984:   Instructor's Handbook on Meteorological  Instrumentation,
    F. V. Brock.  Editor. NCAR Technical Note, NCAR/TN-237+1A.

U.S.Army, 1975:   Part 2, natural environmental factors. Engineering Design
    Handbook,  Environmental Series. Department of the Army, Material Command.

WHO. 1971:   Guide to meteorological instrument and observing practices.  Vorld
    Meteorological Organization No. 8TP3, 4th edition. Geneva', Switzerland.
                        a U.S. GOVERNMENT PRINTING OFFICE: 1994-550-001/80371

-------