ENVIRONMENTAL PROTECTION AGENCY
                  NATIONAL ENFORCEMENT INVESTIGATIONS CENTER
                    P.O. Box 25227  Denver Federal  Center
                          Denver, Colorado 80225
                                  March 1983
          National Enforcement Investigations Center, Denver
CIS. Environmental Protection Agency
                                                    Office of Enforcement

-------
                     ENVIRONMENTAL PROTECTION AGENCY
               NATIONAL ENFORCEMENT INVESTIGATIONS CENTER
                 P.O. Box 25227   Denver Federal Center
                        Denver, Colorado  80225
                                 March 1983
                              GUIDELINES FOR
               ANALYSIS OF VOLATILE ORGANIC COMPOUNDS IN AIR
          A  INTRODUCTION AND GUIDANCE FOR PLANNING FIELD STUDIES

          B  SAMPLE COLLECTION

          C  THERMAL DESORPTION AND GC/MS ANALYSIS

          D  PERMEATION TUBE PREPARATION AND CALIBRATION

          E  TENAX-GC® SAMPLE TRAP PREPARATION AND SCREENING
®  Tenax-GC is a registered trademark of Enka N.V., The Netherlands.

-------
INTRODUCTION AND GUIDANCE FOR PLANNING FIELD STUDIES

-------
This introduction to the NEIC air sampling and analysis procedures provides
background  information  on  air sampling by various means and the reasoning
behind  the  selection of procedures  for use at  NEIC.   Any  environmental  an-
alytical  data  can  be  no better than  the  sampling plan and the protocols
used to collect  samples.   This  is especially true for air sampling  for  or-
ganic compounds.   The objectives of a field study (for example:  great  in-
terest  in one  or a few known compounds) can indicate, if not require, the
use of  modifications to standard sampling protocols.   Stated more explicit-
ly, air sampling and analytical methods need to be tailored to the  object-
ives of the field  study.   Therefore,   is  is important that both  field and
laboratory  personnel understand the basic physical processes and principles
of organic  air pollutant analysis.

Approaches  to Air Sampling

Almost  every organic analysis,  including  organics in air, depends on get-
ting the  sample  or  sample  extract onto some type  of chromatographic column
for the actual  quantisation step.   (This discussion ignores in situ spectro-
scopic  methods such as  infrared because such  methods are limited  in the
number  of compounds which  can be determined and because the potential for
interferences  limits  their application to known  atmospheres.)   The most
straightforward,  and therefore  the  best,  way  to  analyze  air  would be to
place a measured amount of the air directly onto the chromatographic column.
However,  using gas  chromatography  limits  the  sample  size to a few cm3  at
most,  so  that this  direct approach does not provide the necessary sensitiv-
ity in  mos* cases.1  In order to increase sensitivity,  three categories of
methods have been  employed to trap organic compounds from a stream of air
to increase the effective sample size.  The three types of methods  are liq-
uid sorbents,  cryogenic trapping,  and  solid sorbents.

Liquid Sorbents

Liquid  sorbents,  or bubbler  trains, are inconvenient  to  use  both  from  a
field  and laboratory standpoint.  Their use ought to  be limited to special
situations where a  chemical  selectivity can  be achieved  in the  trapping
step which outweighs the disadvantages of  the  technique.

-------
                                                                       A-2
Cryogenic Trapping

Cryogenic trapping  involves  drawing the air sample over a cold inert sur-
face so that when the trap is warmed, the organics are contained in a small
volume.  The  method works best for very volatile compounds (gases at room
temperature)  although  it  has  been  used  for compounds  as  involatile as tri-
chlorobenzene.2  An obvious disadvantage of the method is the need for cry-
ogenic material,  such  as  liquid argon or liquid  oxygen  or dry  ice at the
very least, both  to collect and store the samples before analysis.   Other
problems encountered with cryogenic  trapping are  plugging of the trap with
frozen water  vapor  and reproducibly obtaining an  inert  trapping surface.
The most extensive  use of cryogenic  trapping has  been by a group headed  by
Hanwant Singh.  They perform  analyses in the field using a mobile labora-
tory to avoid problems associated  with storing and shipping samples.   The
analyses are  done by GC using  classical  (not mass  spec)  detectors.  Refer-
ence 3 contains a brief description of their methods and summarizes results
they have obtained  in urban ambient air.

Solid Sorbents

Solid sorbents  are  the most applicable  to general  purpose air sampling be-
cause of their  relative ease of use  and  storage and the  wide range of com-
pounds which  can  be sampled.   Analysis can be performed either by solvent
elution of  the  sorbent or by thermal  desorption.   With thermal  desorption,
the entire  sample is analyzed giving lower detection limits than solvent
elution for a given  sample size.   However, solvent elution nllows reanalysis
of the sample,  which is not possible  with thermal  desorption.   Thermal de-
sorption and  solvent elution  should be thought of as complimentary tech-
niques with a range  of  overlap in their applicability.  Except for special-
ized methods,  compounds compatible  with solvent elution analysis are solids
at room temperature, while thermal desorption can be used for liquids and
more volatile solids.   Compounds which are gases at room temperature can be
analyzed by thermal desorption with a judicious  choice  of solid sorbent.

-------
                                                                        A-3
 Choice  of  Sampling  Technique

 The  method expected to be most applicable to NEIC needs  (for example, air
 sampling  near a hazardous waste site)  is  trapping on Tenax solid sorbent
 with  analysis by thermal  desorption.   In situations where  only relatively
 non-volatile  compounds  such  as pesticides, PCBs, or benzo(a)pyrene are of
 interest,  trapping  on combination polyurethane foam  (PUF)/XAD  resin  car-
 tridges followed by solvent  elution is recommended.  The latter technique
 has  been  successfully applied to  a wide variety of compounds  including or-
 ganophosphorous  pesticides by EPA personnel at HERL-RTP.4

 The  properties of  Tenax  are  described more completely below.  Briefly,
 among the  solid sorbents  available,  Tenax  is applicable to  the widest var-
 eity  of compounds  while  not  collecting large  amounts  of water.   Water
 causes  problems  during  the analysis because thermal desorption includes a
 cryogenic  trapping  step,  and  ice plugs the trap  if enough  water was  col-
 lected  with the air sample.   Other  materials which have been used  as  solid
 sorbents have  fewer of the desirable properties possessed by Tenax.  Nearly
 all adsorb more water.   Some  do not have  the  thermal  stability of Tenax
 and others, such as charcoal, can  irreversibly adsorb compounds  limiting
 their range of applicability.

 Properties of Tenax-GC
The structure of Tenax is shown below.
                     -0-/V-

-------
                                                                        A-4
 Tenax-GC  is poly-p-2,6-diphenylphenyleneoxide,  a porous  linear  polymer
 first  used as  a packing material  for gas chromatography.   Several  proper-
 ties  of Tenax  make it suitable for collection of organic volatiles in air.
 They are:

 1-   Excellent thermal stability.   Thermal  analysis methods  (Differential
     Scanning  Calorimetry)  have  been  used to  confirm the stability of
     Tenax.  Breakdown  does not occur  below temperatures of 400°  C.5  This
     characteristic  makes  thermal  desorption feasible and  temperatures as
     high  as 400°  C. will   not  increase background due to breakdown of the
     polymer.

 2.   Quantitative desorption.   Tenax does  not exhibit losses due to irre-
     versible adsorption as charcoal  does.6  Decomposition of sorbates on
     charcoal has been reported.7

 3-   Low background  contamination.  Virgin  Tenax is Sohxlet extracted for
     ==18 hours with  methanol.   It  is then thermally  conditioned at  an ele-
     vated  temperature  (275-325°  C.)  for  20 to  30  minutes.   Traps are
     easily  rebaked  and returned  to  a low background  state for reuse.
     After  solvent extraction  and  thermal conditioning,  Tenax  shows  negli-
     gible  background contribution.  However, the sampling  process  exposes
     any solid sorbent to ozone and other oxidants which can be expected to
     cause  artifacts.  Artifacts  reported  from the thermal  decomposition
     and/or reactions  of Tenax  include ethylene oxide7, alkylbenzenes8,
     styrene8,  benzene8, alkylphenols8, acetophenone9,  and benzaldehyde9.

     In our experience  at  NEIC, only acetophenone  and  benzaldehyde have
     been observed as artifacts in upwind  field samples.

4.   High collection  efficiency.  Tenax collection  efficiency  of  100% has
     been reported with a wide variety of compounds10 (when breakthrough
     volumes are  not  exceeded).  Testing has also been conducted under com-
     monly encountered analytical  conditions.   Results  indicate that col-
     lection efficiency does not drop with  repeated  thermal  desorption10'll

-------
                                                                        A-5
     and  that  breakthrough  volumes  are  somewhat dependent on  relative  humi-
     dity12  and  C02 concentration12, even though  no  apparent decrease in
     collection  efficiency  is observed with  relative  humidity10/13.   (Col-
     lection efficiency  and breakthrough volume are  not  the  same thing.)

5.   Low  affinity for water.  Water vapor is poorly retained  on Tenax14/ 1S,
     a  distinct  advantage.   In addition to the plugging problem mentioned
     previously, adsorbed water presents a potential  medium for hydrolysis
     reactions and  for  collecting potentially reactive gases such as  NO
     and  S0215.

Limitations of Tenax

Although  Tenax has  many desirable  properties,  it  does have  limitations.
Because of  these limitations,  it  may be desirable  to  use other sorbents in
conjunction  with Tenax.   Materials  such as charcoal8,16  and Ambersorb
      ®
XE-340  8 retain very volatile compounds much better than Tenax.  Ambersorb
XE-340  is essentially carbonized  XAD resin which has  the desirable adsorp-
tive properties  of charcoal, hopefully  without  the  undesirable features
such as active sites  due to metals.   If it  is necessary to  sample  very
volatile  compounds, Ambersorb XE-340 can be used by itself  or behind Tenax
traps.

Breakthrough Volumes

Two terms commonly  used  to  describe collection efficiency in  air are Elu-
tion Volume and Breakthrough Volume.  Due to  ambiguity by early researchers,
these terms were often  used interchangeably?;10.   To  clarify  the meaning,
they are defined below.

     Elution Volume  - The volume  of air sampled which is  required to move
     the mass transfer zone  to the end of the  available packing bed.
     Ambersorb XE-340  and  XAD are registered trademarks  of  Rohm & Haas
     Company.

-------
                                                                       A-6
     Breakthrough Volume  -  That  volume of  air which  purged  50% of the
     adsorbed vapor out of the cartridge.

The two definitions can be represented diagramatically.

                                 FIGURE 1
                              ELUTION PROFILE
Concentration
at exit of trap
                              Volume Sampled

     V = point  where  actual  breakthrough of the sorbate  begins  (Elution
         Volume)
     VB = 50% breakthrough (Breakthrough Volume)

Another way of visualizing this concept is shown  in Figure 2.
           100%
 % of Challenge
 Concentration
 at Trap Exit
                                 FIGURE 2
                              FRONTAL PROFILE
50X Breakthrouah
                                    Elution Volume
                              Volume Sampled

-------
                                                                           .A-7
 Elution analysis entails injecting a  small  quantity of adsorbate onto  a
 cartridge  in  a  very  small time.   During  frontal  analysis,  the  sample  injec-
 tion  is continuous.

 In  actuality, the elution volume  and  the 50% breakthrough volume may be
 very  similar  for compounds with  low breakthrough  volumes.  Compounds with
 high  breakthrough volumes will have a  larger difference between the two
 values.
 Superimposing  Figures  1 and 2 will  show the relationship between frontal
 breakthrough (Figure 2) and an elution peak  (Figure I)11.
                                           Frontal  Profile
Concentration
                                                                 Figure 3
                                       -p  V \J I UUIC
               [_ 50% Breakth'rough_   (V  Specie Restitution  (Nation}
                 Elution    ,
               I   Volume ~~~"
Factors Affecting Breakthrough Volume

Factors which affect or may affect breakthrough volume include temperature,
humidity, chemical  class  of the sampled compound, C02 concentration,  the
total organic concentration  in  the sampled atmosphere or  the  concentration
of particular compounds, and changes  in the sorbent surface caused by reac-
tions with ozone  or other oxidizing  species.   All porous polymer sorbents

-------
                                                                        A-8
 work by allowing airborne compounds to diffuse into pores within the poly-
 mer.   This process requires a  finite  time so that collection  efficiency
 will  eventually  decrease  as  the sample  flow  rate  in increased.17   This  prob-
 lem  would probably never be encountered  because  the  flow rates are  higher
 than are  typically used or even possible  with  common  pumping  systems.   How-
 ever,  one should be aware of  the possibility.  The  dependence of break-
 through volume on temperature  is well  known and  has  a  theoretical basis16'
 *? \  extrapolation  of breakthrough volumes  to other temperatures is an often
 used and  accepted procedure.   Tabulated breakthrough  volumes  are  useful in
 estimating relative detection limits.

 The  dependence on relative humidity and  C02 concentration has  been  deter-
 mined  empirically for selected compounds on  Tenax.12   The breakthrough
 volumes of four  compounds decreased 22 to  43%  upon going  from 0 to 87%  rel-
 ative  humidity in a laboratory  situation.  When C02  was  added  in addition
 to water,  breakthrough volumes  decreased by approximately an additional 25%.

 C. R. McMillin and co-workers sampled indoor and outdoor  air using Tenax  as
 the  first stage  of a three-sorbent  trapping  system.8  They found  great  and
 unexplained differences in  the  indoor and outdoor breakthrough volumes on
 Tenax  (other  sorbents were not  tested),  implying  that breakthrough volumes
 determined  in  the  laboratory cannot be  used  in quantitating field results.
 These differences  are the observable effects of the many factors  affecting
 breakthrough volume.

The  only  practical approach to  obtaining  quantitative  field  data is tc
 avoid breakthrough  by limiting  sample size and to include tests for  break-
 through in the quality assurance plan so that you are aware if breakthrough
did  occur.  Tabulated breakthrc igh  volumes or past experience can be used
 to set sample size, but should not be used for quantisation.

Generic Quality Assurance  Plan for Air Sampling

The  procedures  listed are recommended as  a  general approach  to any air
sampling study.  The reasons  for each procedure are  also listed.

-------
                                                                        A-9
 1.    Keep  the sampling time to the minimum which will give adequate sample
      volume.   This point depends on  the  objectives  of  the  study.   Presum-
      ably, you are trying to  detect contributions  from  a particular source.
      The shorter  the  sampling time, the more  stable  weather  (wind speed and
      direction) conditions  will  be during the  sampling, resulting in better
      directional  resolution in the data.   An  obvious corollary  is  to col-
      lect all  samples  as nearly  simultaneously as  possible.

 2.    Do  not  use  a  sample volume  larger than necessary  to give the  desired
      detection limits.  During sampling,  both  the  solid sorbent  and adsorbed
      compounds are exposed  to ozone and other reactive  species.   Reactions
      between  adsorbed compounds  are  also possible.   These possibilities
      cannot be avoided, but minimizing the sample  size also minimized their
      effects.

 3.    Take all  samples  using tandem (connected  in series) sorbent tubes with
      a sample  size chosen  to avoid breakthrough  on  the first tube.  This
      procedure is  based  on  the assumption that quantitative data for more
      volatile  compounds are given higher  priority  than lower detection lim-
      its for  less  volatile  compounds.  Avoiding breakthrough is necessary
      to obtain quantitative data; analyzing tandem tubes provides a check
      for breakthrough.

4.   Always take  duplicate  samples  downwind of the  source  as  a minimum;
     triplicate samples are recommended.   The downwind samples are the most
     important samples because they are most likely to show positive results;
     the upwind sample is  the next most  important because  it shows back-
     ground levels.  Taking duplicate  samples downwind helps to ensure that
     data will be  available for  the most  important sampling point « ven if
     one laboratory analysis  fails.   (Using thermal  desorption, there is
     only one chance at the analysis.)  The duplicate or triplicate  samples
     also serve as a check  of sampling and analysis precision.

5.   Take duplicate breakthrough  spikes upwind of the source in addition to
     the upwind sample(s).   A breakthrough spike consists of a tandem pair

-------
                                                                       A-10
     of  tubes  with the first tube  spiked with  known  amounts  of target com-
     pounds.   The  tandem pair is then sampled over in the normal manner.  A
     breakthrough  spike  is  the  most valid matrix spike which  is practical
     in  most  air sampling projects.   The results obtained from the break-
     through  spike are the  best performance that can be expected from the
     sampling  and  analysis procedure.  If a spiked compound is not retained
     on  a  tube during  sampling,  it cannot be expected to be  sampled  relia-
     bly.   The duplicate  breakthrough  spikes  are a check of the precision
     and accuracy  of the methodology.

6.   Carry  triplicate  field spikes to the field.  The field spikes are sam-
     ple tubes which are spiked with target compounds but not sampled over.
     The field spikes  serve  to check for losses of target compounds  during
     shipment  and  storage and as a check of analytical  precision.   At the
     time the  spiking  for the breakthrough and field spikes is performed,
     an  additional  spike  is  performed  which is stored  in  the laboratory
     under  the  best possible  conditions.  This  spike serves as a reference
     for the field and duplicate spikes.

7.   Include a  field blank  in each container  of sample  tubes to check for
     contamination during storage or shipment.

8.   One can expect artifacts from any  sampling technique.   Be aware of the
     artifacts which can be  expected with the  methodology employed.

9.   Know the  principles  involved  in air sampling methodologies and avoid
     situations which  might  cause  artifacts or errors.   For example, two
     situations to  avoid would be exposing sample traps to high concentra-
     tions  of  ,rganic  vapor such as  gasoline fumes, and exposing the traps
     to high temperatures during shipment or storage.

-------
                                                                            A-11
                            FOOTNOTES  AND  REFERENCES


 1.   Tom  Spittler and co-workers  in U.S.  Environmental  Protection Agency
     Region  I, Boston,  have very  successfully  employed  this  direct approach
     to monitor  solvent-type  compounds.

 2.   H. B. Singh, L. J. Salas, A. J.  Smith, and  H. Shigeishi, Atmos.
     Environ., 15, 601  (1981).

 3.   H. B. Singh, L. J. Salas, and R. E.  Stiles, Environ. Sci.  Technol.,
     16,  872  (1982).

 4.   R. G. Lewis and K. E.  MacLeod, Ana2. Chem., 54, 310  (1982).

 5.   "Selection  and Evaluation of Sorbent Resins for the  Collection of
     Organic Compounds", A. D. Little,  Inc., EPA/600/7-77/044,  April  1977.

 6.   E. 0. Pellizzari,  B. H.  Carpenter, J. E.  Bunch, and  E.  Sawicki,
     Environmental Science  and Technology, 9,  556  (1975).

 7.   "neveJopment of Method for Carcinogenic Vapor Analysis  in  Ambient
     Atmospheres", Research Triangle  Institute,  EPA-650/2-74-121,  July  1974.

 8.   "Potential Atmospheric Carcinogens,  Phase 2/3, Analytical  Technique
     and Field Evaluation", Monsanto  Research  Corp., EPA-600/2-81-106,
     June 1981.

 9.   "Artifact Problems in  Atmospheric Analysis of Organic Compounds  and
     Strategies for Minimization," R. E.  Sievers,  presented  at  National
     Symposium on Monitoring  Hazardous Organic Pollutants in Air,  Raleigh,
     N.C., April 28 to  May  1, 1981.    Also by NEIC  experience.

 10.  "Development of Analytical Techniques for Measuring Ambient Atmospheric
     Carcinogenic Vapors",  Research Triangle Institute, EPA-600/2-75-076,
     November, 1975.

 11.  "Characterization of Sorbent Resins  for Use in Environmental  Sampling",
     Research Triangle  Institute, EPA-600/7-78-054, March 1978.

 12.  "Further Characterization of Sorbents for Environmental Sampling",
     A. D. Little, Inc., EPA-600/7-79-216, September 1979.

 13.  E. D. Pellizzari,  J.  E.  Bunch,  R. E.  Berkley, J.  McRae, Analytical
     Letters, 9, 45 (1976).

14.  Dravnieks, et al., Environmental Science and Technology, 5, 1220
     (1971).                                               **  -'

15.  "Analysis of Organic Air Pollutants by Gas Chromatography and Mass Spec-
     troscopy, Final  Report",  Research Triangle Institute, EPA-600/2-79-057,
     March 1979.

16.  K. J. Krost, E.  D.  Pellizzari,  S. G.  Walburn, and S.  A.  Hubbard,
     Ana2. Chem., 54,  810 (1982).

17.  "Characterization  of SorJbent Resins for Environmental Sampling," A. D.
     Little,  Inc.,  EPA-600/7-78-054,  March 1978.

-------
B    SAMPLE COLLECTION

-------
                     ENVIRONMENTAL PROTECTION AGENCY
                NATIONAL ENFORCEMENT INVESTIGATION CENTER
                     Box 25227  Denver Federal Center
                          Denver, Colorado  80225


                  Volatile Organic Air Pollutant Analysis
                             Sample Collection
                                March 1983

1.0  Introduction

     1.1  This procedure describes the collection of air samples on sampling
          tubes containing a solid sorbent.  Tenax-GC  is the most commonly
          used sorbent.  Adsorbed organic compounds are detected by thermal
          desorption of the sorbent onto a gas chromatograph column for
          GC/MS analysis.   The procedure presented here was designed to per-
          form reasonably well on a wide range of compounds.  In general,
          those organic compounds that are liquids at room temperature are
          well suited for analysis using this method and Tenax as an adsor-
          bent.  If only compounds of a narrow volatility range are of in-
          terest, it is probable that the sample size and/or sorbent mater-
          ial could be changed to yield superior performance for the com-
          pounds of interest.   For example, compounds as low in volatility
          as benzo(a)pyrene have been analyzed by similar procedures.   Al-
          though this procedure was specifically designed for the use of
          Tenax sorbent, other sorbents can be used.

2.0  Limitations

     2.1  The sample traps are essentially short chromatographic columns.
          Retention of chemicals is dependent upon adsorption characteris-
          tics of the chemical/resin system.   Factors influencing retention
          include:   temperature, flow rate, air volume, vapor pressure of
          the chemical, and sample matrix.   Volatile species like vinyl
          chloride are only moderately retained while other chemicals like
          chlorobenzene are retained very well.   All  chemicals will experi-
          ence breakthrough under the correct conditions.   Table I lists
          breakthrough volumes for some relevant chemicals.   The volumes
          represent the amount of air sampled when 50% of the collected
          chemical  is lost through the trap.   This data was compiled by
          Pellizzari  in Reference 9.7.   Data for chemicals where the sample
          volume exceeds the breakthrough volume represent minimum concen-
          trations.   The data  in Table 1 can be used to estimate an appro-
          priate sample size.   When doing this,  one must decrease the vol-
          umes shown in Table  1 by the factor §f because the NEIC sample
          traps contain 0.8g of Tenax rather than 2.2g.   In general, the
          sample size should be 25 liters or less.
   Registered Trademark of Enka  N.V.,  The  Netherlands;  appears  hereafter
   without the ®.

-------
                                                                                                                       B-2
 Chemical Class
 •~^^^—^^«.^



 Alcohols




 Aldehydes


 Amines
 Aromatics
 Esters
Ethers
Halogenated
Ethers

Halogenated
hydrocarbon
                                                           Table  1

                       Tenax GC Breakthrough Volumes  for Selected Compounds at Various Temperatures
                                               Data Taken From Reference 9.7
      Compound
  Methanol
  n-Propanol
  Allyl alcohol

  Acetaldehyde
  Benzaldehyde

  Dimethyl amine
  Isobutylamine
  t-Butylamine
  Di-(n-buty1)amine

  Pyridine
  Aniline

  Benzene
  Toluene
  Ethyl benzene
  Cumene

  Ethyl acetate
 Methyl acrylate
 Methyl methacrylate

 Diethyl ether
 Propylene oxide

 2-Chloroethyl ethyl ether
 Bis-(chloromethyl)ether

 Methyl chloride
 Methy.l bromide
 Vinyl  chloride

 Methylene chloride
 Chloroform
 Carbon tetrachloride

 1,2-Dichloroethane
 1,1,1-Trichloroethane
 Tetrachloroethylene
 Trichloroethylene

 l-Chloro-2-methylpropene
 3-Chloro-2-methylpropene

 1,2-Dichloropropane
 1,3-Dichloropropane
 Epichlorohydnn (1-chloro-
  2,3-epoxypropane)

 3-Chloro-l-butene
Allyl chloride
4-Chloro-l-butene
 l-Chloro-2-butene
^ — _
b p (°
64.7
97.4
97
20
179
7.4
69
89
159
115
184
80 1
110.6
136.2
152.4
77
80
100
34.6
35
108

-24
3.5
13
41
61
77
83
75
121
87
68
72
95
121
116
64
45
75
84


C)
1
27
32
3
7,586
9
71
6
9,506
378
8,128
108
494
1,393
3,076
162
164
736
29
13
468
995
8
3
2
11
42
34
53
23
361
90
26
29
229
348
200
19
21
47
146

60
Temperature (°F)
70
80
Breakthrough Volumes for 2.7 q
1
20
23
2
5,152
6
47
5
7,096
267
5,559
77
348
984
2,163
108
111
484
21
9
336
674
6
2
1.5
9
31
27
41
18
267
67
20
22
162
253
144
15
16
36
106
0.8
14
16
2
3,507
4
34
4
4,775
189
3,793
54
245
693
1,525
72
75
318
15
7
241
456
5
2
1.25
7
24
21
31
15
196
50
16
17
115
184
104
12
12
27
77
0.6
10
11
1
2,382
3
23
3
3,105
134
2,588
38
173
487
1,067
48
50
209
11
5
234
309
4
1
1.0
5
18
16
23
12
144
38
12
13
81
134
74
9
9
20
56
90
100
Tenax (liters)
0.4
7
8
0.9
1,622
2
16
2
2,168
95
1,766
27
122
344
750
32
34
137
3
4
124
209
3
1
0.8
4
13
13
18
9
106
28
9
10
58
97
54
7
6
15
40
0 3
6
0.7
1,101

11
1
1,462
67
1,205
19
86
243
527
22
23
90

3
89
142
2.5
0.9
0.6
3
10
10
14
j
78
21
7
i
a
41
70
39


12
29

-------
                                                                                               B-3
                                   Table 1 (cont.)
Tenax GC Breakthrough Volumes  for Selected Compounds at Various Temperatures
                        Data Taken From Reference 9.7
Chemical Class

Halogenated
hydrocarbon
(cont. )

Hydrocarbons




Inorganic gases



Ke tones

Nitrogenous
hydrocarbons
Oxygenated
hydrocarbons


Sulfur
Compounds
Compound

Chlorobenzene
o-Dichlorobenzene
m-Dichlorobenzene
Benzyl chloride
Bromoform
Ethyl ene di bromide
Bromobenzene
n-Hexane
n-Heptane
1-Hexene
1-Heptene
2,2-Dimethylbutane
2,4-Diemthylpentane
4-Methyl-l-pentene
Cyclohexane
Nitric oxide
Nitrogen dioxide
Chlorine
Sulfur dioxide
Water
Acetone
Methyl ethyl ketone
Methyl vinyl ketone
Acetophenone
Ni tromethane
Aniline
Acrolein
Glycidaldehyde
Propylene oxide
Butadiene diepoxide
Cyclohexene oxide
Styrene oxide
Phenol
Acetopheonone
b-Propiolactone
Oiethyl sulfate
Ethyl methane sulfate

b.p. ('
132
181
173
179
149
131
155
68 7
98.4
63.5
93.6
49.7
80.5
53.8
80.7
_
-
.
100
56
80-2
81
202
101
184
53
34
132
194
183
202
57
208
86

50

60
Temperature (°F)
70
°C) Breakthrouah Volumes
899
1,531
2,393
2,792
507
348
2,144
32
143
28
286
0.5
435
14
49
0
0
0
0.06
0.06
25
82
84
. 5,346
45
3.864
19
364
35
1,426
2,339
5,370
2,071
3.191
721
40
5,093
653
1,153
1,758
2,061
386
255
1,521
23
104
20
196
0.4
252
10
36
o
o
o
0.05
0 05
17
57
58
3,855
34
2.831
14
247
24
1,009
1,644
3,926
1,490
2,382
514
29
3,681
473
867
1,291
1,520
294
188
1,079
]?
75
15
135
0.3
146
8
26
o
o
o
0.03
0 04
12
39
40
2,767
25
2,075
10
168
17
714
1,153
2,870
1,072
1,778
366
21
2.564
80
90
100
for 2 2 q Tenax (liters)
344
656
948
1,125
224
138
764
17
1 £
55
11
93
0.2
84
6
19

o
g
0 02
0 03

27
28
2,000
19
1,520
8
114
11
506
811
2,094
769
1,327
261
15
1,914
249
494
697
830
171
101
542

39
8
64
0.2
49
4
14



0.02
0.01

19
19
1,439
14
1,114
6
77
8
358
570
1,531
554
991
186
11
1,384
181
372
510
612
131
74
384

29
6
44
0.1
28
3
10

0

0
0 01
0

13
14
1,037
11
817
4
52
5
253
400
1,119
398
740
132
8
998

-------
                                                                           B-4
     2.2  The accuracy of the data produced from the analysis of samples ob-
          tained using this procedure depends on the care with which samp-
          ling is performed.  Particular attention must be given to the cal-
          ibration of pumps, checking to demonstrate that the sampling rate
          was constant, and to the handling of sample tubes to avoid contam-
          ination.   The sampling tubes should be kept in sealed culture
          tubes except for the time required for set up and sampling.   The
          tubes should never be handled without using nylon gloves or tis-
          sues to prevent contamination by body oils.

     2.3  The data in Table 1 show that the effects of temperature on sample
          breakthrough volume are significant.   For many of the compounds
          listed in Table 1, the breakthrough volume at 90° F is only 20 to
          25% of the breakthrough volume at 50° F.

     2.4  In order to check for sample breakthrough, each sample is taken
          using a tandem tube arrangement.   If a particular compound is de-
          tected on the first tube but none is seen on the second tube,
          then that compound did not experience breakthrough.

3.0  Equipment

     3.1  Sampler - DuPont model P4000 or equivalent personnel sampler.
          Capable of adjusting and monitoring the flow over the range of
          0.1 to 1  liter per minute (2pm) with a trap in place.

     3.2  Mass flow meter - Portable unit equipped with a teflon fitting to
          measure the flow through a sampling trap.   It should have a range
          of 0-2 £pm.

     3.3  Sample traps - Glass sampling traps packed with the selected sor-
          bent.   See the procedure "Tenax Sample Trap Preparation and
          Screening".

     3.4  Sampling line - 2-5 feet of 1/4"  o.d.  tygon tubing with a teflon
          fitting at one end to attach to the sampling traps.

     3.5  Swagelok - 5/8" union.

4.0  Calibration Procedure

     4.1  A mass-flow meter is used in line between the pump and the adsor-
          bent traps to calibrate the pump  before san >ling begins at each
          station.   The pump is rechecked with the mass-flow meter after
          the sampling period is complete.

5.0  Sample Collection

     5.1  Sample tubes are packaged inside  screw cap culture tubes placed
          in metal  cans with a compression  fit closure for transport to the
          field.   The NEIC employs virgin paint cans for packaging.  Using
          a clean tissue or wearing a nylon cloth glove, remove a numbered
          sample trap from its culture tube,  and reseal  the culture tube.

-------
                                                                           B-5
     5.2  Inspect the trap for damage such as broken glass, loose glass wool
          plugs, or spilled resin.  If the trap is damaged, replace in the
          culture tube and return to the laboratory unused.

     5.3  Attach the tandem traps to the calibrated sampling pump.  See
          Figure 1.

     5.4  Begin sampling, noting the start time and sample pump flow meter
          reading.   Select sample volumes so as to avoid breakthrough of
          target pollutants from the first trap to the second.   For most
          purposes a 25-liter sample collected at 0.5-1.0 liters per minute
          is desirable.

     5.5  Record the weather conditions occurring during sampling including
          temperature, wind speed and direction, humidity, and barometric
          pressure.

     5.6  Stop sampling, noting the end time and sample pump flow meter
          reading.   Replace the trap into a culture tube.   Reseal with the
          teflon-lined septum cap and tag.   Note on the tag the trap number
          and whether the trap was the front or back of the tandem pair.
          The front trap is the one sampled air passes through first.

     5.7  Return culture tubes containing sample traps to the paint can
          and reseal  the can.   Be sure to tag the "field blank" sample in
          each can and any field spikes (will be total of at least three).

6.0  Quality Control

     6.1  Sample pumps are calibrated daily.   During sampling any flow rate
          changes are noted by monitoring the flow meter on the sampler.
          Changes in  flow up to 10% are acceptable.   If the change is  greater
          than 5%,  the beginning and ending flow rates are averaged to give
          the flow rate.

     6.2  Triplicate  samples indicate the reproducibility of the overall
          sampling  and analysis.   Triplicate  samples will  be collected at
          least at  one sampling station.   The triplicate sample station
          (or one of  the duplicate sample locations) should be  that station
          most directly  downwind of the source being sampled.   The tripli-
          cates should be collected at the  same place,  at the  same flow
          rate and  at the same time.   It is very important to  have the sam-
          pling as  identical  as physically  possible.

     6.3  Breakthrough spikes  give an  indication of  which  compounds  would
          have broken through  under the field conditions of the sampling.
          A  breakthrough spike [Figure 2] consists of a tandem  pair of
          traps,  the  front trap having been spiked with a  standard set of
          compounds in the laboratory.   A sample is  then collected in  the
          normal  manner  using  the tandem traps.   Duplicate breakthrough
          spikes  should  be sampled at  the same time  and rate as the  regular
          field sample at the  sample  station  most directly upwind of the
          source  being sampled.

-------
                                          FIGURE 1:   Tandem Air Sample Traps
                                                   FLOW
To Pump
— 1
(\ • 1


•» 	 1
	 . — j j 	 ;N x^~ '•• '•' ' ••' " ' • • '••••• i < )
Ijl I"""! fc'';';::.:, •''-:. '•- :^: N I
>~j •* — ^^ 	 ' ••'-.' -•-. • h i
1/4" Teflon Union Air Sample Trap \L F

- h
i
i
'• X
/ :_.'•;,--.'. - ^
':'•' Z. '•''•'•• s ' ''''•.

Air Sample Trap
                                                         5/8" Brass Union
                                                         with Teflon ferrules
                                         Figure  1.  Tandem Air Sample  Traps
                                                                                                                       CO

                                                                                                                       CTl

-------
                                        FIGURE 2:   Breakthrough Spike
                                                    FLOW
To pump
      1
      1/4" Teflon Union
*'
!

!?fe ; ,".'.':'•• ,' • :> I
J&¥ '• .,'' >';.'.•/./';. '•" ;?--; 1 i
^*~*1 * ' 	 ',.. ,..,... j »
Unspiked Sample Trap (j
I
!
)
i
!
. I
„, „ , 1
\ 1


.


	
r*
	 i
)
i

t
' S
\
l_
••:' -. . ' •:-.- ;. . • '^
•'•" •- "- ', ; -', - "---s -. - ('- ,
-v • •*.•"•.. • -. -:-y
Soiked Samole Trao
                                                         5/8" Brass Union with
                                                           Teflon ferrules
                                           Figure 2.  Breakthrough Spike
                                                                                                                       CD
                                                                                                                       I

-------
                                                                           B-8
     6.4  Contamination in each sample transport container (paint can) is
          monitored by a "field blank".   A field blank is a sample trap
          which is not sampled.  It is transported and stored the same as
          the samples.  Field personnel  designate the trap to be used as
          the field blank.

     6.5  Deterioration of the samples is monitored by a "field spike".   A
          field spike is a sample trap which is spiked in the laboratory
          prior to going out to the field.   The field spike is stored and
          transported alongside the samples.

     6.6  At the time laboratory personnel  prepare the field and breakthrough
          spikes,  they will  prepare a reference spike which remains in the
          laboratory.

     6.7  Samples  can be stored in a dark,  organic vapor-free area at -20° C
          for up to four weeks before analysis according to Reference 9.10.
7.0  Options
     7.1   In the event of unknown atmospheres suspected of containing high
          levels of contaminants, two samples could be collected,  one at
          the normal  sampling rate,  and another at one tenth the normal
          rate.

     7.2   If specific compounds  are  of special  interest,  flow rates  and
          sampling times may be  changed (e.g.,  a compound with a high break-
          through volume, suspected  in low concentrations might be sampled
          at 1  liter per min for 100 minutes).

     7.3   If particulate matter  may  provide an  unwanted contribution to  sam-
          pled  organics, filters -are available  which will prevent  particu-
          late  greater than 0.5  urn in size from reaching the sorbent traps
          (and  thus being thermally  desorbed when the trap is analyzed).
          Any filter must be used with the realization that organic  com-
          pounds may be stripped from particulates on a filter by  the sam-
          pled  air flow, so that a total  elimination of the contribution
          of organics from particulate is not possible.   However,  compounds
          stripped from particulates by the sampling process would have  to
          be considered readily  available for volatization, and would prob-
          ably  be of interest.

          7.3.1   The  filters used are sold commercially as filters for liq-
                 uid  chromatography  solvents.   The filters are Millex -SR,
                 Millipore Corp., Bedford,  MA,  Catalog #SLSR025NS, 0.5 urn
                 PTFE (polytetrafluoroethene).   The PTFE  filter itself is
                 encased inside  a hard plastic  holder equipped with  leur
                 fittings.   Other filter  pore sizes are available.

          7.3.2   The  filters can be  connected to sample traps by forcing  one
                 end  of the plastic  case  directly into the quarter-inch  end
                 of the glass trap.

-------
                                                                           B-9
8.0  Sample Analysis

     8.1  Samples are analyzed by the procedure "Thermal Desorption and
          GC/NS Analysis of Air Samples" (see Section C).

9.0  References

     9.1  Bertsch, Wolfgang, Chang,  Ray C.  and Albert Zlatkis,  'The Determi-
          nation of Organic Volatiles in Air Pollution Studies:   Characteri-
          zation of Profiles", Journal  of Chromatographic Science.  Vol.  12,
          pp 175-182, April 1974.

     9.2  Pellizzari, Edo D.,  "Development of Method for Carcinogenic Vapor
          Analysis in Ambient Atmospheres", EPA-650/2-74-121, July  1974.

     9.3  Pellizzari, Edo D.,  Bunch,  John E. , and Ben H.  Carpenter, "Collec-
          tion and Analysis of Trace  Organic Vapor Pollutants in Ambient
          Atmospheres:  Technique for Evaluating Concentration  of Vapors by
          Sorbent Media", Environmental Science and Technology.  Vol 9,
          pp 552-553, 1975.

     9.4  Pellizzari, Edo D.,  "Development of Analytical Techniques for Mea-
          suring Ambient Atmospheric  Carcinogenic Vapors",  EPA  600/2-75-076,
          November 1975.

     9.5  Pellizzari, Edo 0.,  'The Measurement of Carcinogenic  Vapors in
          Ambient Atmospheres", EPA  600/7-77-055, June 1977.

     9.6  Pellizzari, Edo D.,  "Analysis of Organic Air Pollutants by Gas
          Chromatography and Mass Spectroscopy:  Final Report",  EPA
          600/2-79-057,  March 1979.

     9.7  Pellizzari, Edo D.,  "Ambient Air Carcinogenic Vapors:   Improved
          Sampling and Analytical Techniques and Field Studies",  EPA
          60012-79-081,  May 1979.

     9.8  "Volatile Organic Air Pollutant Analysis - Permeation  Tube Prepa-
          ration and Calibration", NEIC,  March 1983.

     9.9  "Volatile Organic Air Pollutant Analysis - Tenax Trap  Preparation
          and GC Screening", NEIC, March 1983.

    9.10  Pellizzari, Edo D.,  "Analytical Protocol:   Personal Monitoring of
          Vapor Phase Organic Compounds in Ambient Air (RTl)".

-------
THERMAL DESORPTION AND GC/MS ANALYSIS

-------
                        ENVIRONMENTAL PROTECTION AGENCY
                 NATIONAL  ENFORCEMENT  INVESTIGATION CENTER
                     Box 25227, Denver, Colorado  80225
                   Volatile Organic Air Pollutant Analysis
                Using Tenax GC  , Thermal Desorption and GC/MS
                          Analysis of Air Samples
                                March 1983
 1.0   Introduction

      This method describes the GC/MS analysis of the organic components
      of air samples collected on Tenax traps.  The analysis depends on
      proper procedures for the preparation of Tenax traps, and for the
      collection of samples.  Those procedures are documented as the NEIC
      methods "Volatile Organic Air Pollutant Analysis - Tenax Trap Pre-
      paration and GC Screening" and "Volatile Organics Air Pollutant
      Analysis - Sample Collection".

 2.0   Summary of Method

      Samples are collected by drawing a known volume of air through an
      adsorbent resin which traps organic components.  The resin traps
      are analyzed by thermal desorption of the organics into a cryogenic
      trap which is subsequently flash-heated to transfer the compounds
      onto a GC column for GC/MS analysis.

 3.0   Detection Limits

      Detection limits for air samples depend on sample size, retention
      characteristics on Tenax, and the individual sample matrix, among
      other things, but can generally be expected to be in the range of 5
      to 50 |jg/meter3 for a 25-liter sample size.   Table 1 lists minimum
      amounts of representative compounds detectable by this thermal de-
      sorption/GC/MS procedure.

 4.0   Limitations

      4.1  Often, standard reference materials are not available and only
          tentative identifications of unknowns can be achieved.

      4.2  Because of the long time required to prepare accurate permea-
          tion tube quantisation standards, a limited number of chemi-
          cals can be quantitated.

      4.3  Quantisation may not be possible if breakthrough occurs during
          sampling.   The sample traps are essentially short chromatograph-
          ic columns.   Retention of chemicals is  dependent upon adsorption
®  Registered Trademark of Enka N.V.,  The Netherlands; appears hereafter
   without.

-------
                                                                                C-2
                                        Table 1
           GC/MS RESPONSE FACTORS AND DETECTION LIMITS FOR SELECTED COMPOUNDS



Compound
Hexafluorobenzene (IS)
ds-bromoethane (IS)
Bromochlorome thane (SU)
Bromopentafluorobenzene (SU)
Benzene
Carbon Tetrachloride
Chlorobenzene
1,2,4-tn'chlorobenzene
1,2-dichloroethane
1,1,1-trichloroethane
1,1-dichloroethane
1,1,2-trichloroethane
1,1,2,2-tetrachloroethane
Bis(2-chloroethyl)ether
Chloroform
1,1-dichloroethane
1,2-transdichloroethylene
1,2-dichloropropane
Ethyl benzene
Bromoform
Tetrachloroethene
Toluene
Acetone
Hexane
Trichlorofluoromethane
n-Octane
2-Chloro toluene


Spiking
Level (ng)
2500
2500
380
1000
630
1000
400
100
900
270
620
150
160
75
1600
2000
4000
270
160
100
990
320
1000
510
1600
50
70
* ?£RSD — Percent Relative Standard nt*iria+'
*Q*»**** * WA h«^J J l^ CV^AatvAVC tj ^CtJlUQ J U W Y ^CK C*.

Average
Response
Factor
1.000
0.867
2.494
2.101
3.668
0.890
3.146
0.959
1.056
1.333
2.431
1.489
1.694
1.548
1.273
1.280
0.644
1.816
5.024
0.643
1.114
4.694
1.076
0.548
1.139
6.549
4.595
.. 	 STD DEV _ .
Average



%RSD*
.
6.4
16.3
15.7
16.4
14.2
18.3
29.1
17.5
18.2
18.3
20.5
26.4
23.8
23.8
16.4
17.8
18.0
15.0
16.2
17.7
16.6
21.4
12.1
15.3
19.6
20.7
Jnn°s
I Vl/^)
Lower
Limit of
Detection
(ng)

_
200
300
200
400
100
NA**
300
100
200
50
200
20
800
600
1000
60
20
NA
300
100
700
90
500
NA
NA


**  Not Available

-------
                                                                           C-3
          characteristics of the chemical  and the resin system.   Some fac-
          tors influencing retention include:   temperature, flow rate, vol-
          ume of air sampled, vapor pressure of the chemical, chemical class,
          presence of other chemicals, and batch-to-batch variation in the
          sorbent resin.   Due to the many  variables involved, predicting
          breakthrough is very difficult.   Tandem tubes and breakthrough
          spikes are used in the field to  determine if breakthrough has
          occurred.   The sample size should be chosen to try to  avoid break-
          through.

          The breakthrough volume is defined as the amount of air which
          causes 50% of the collected chemical to be lost through the trap.
          Table 1 of the procedure "Sample Collection" lists breakthrough
          volumes compiled by Pellizzari  in Ref.  12.8.  The temperature was
          the only factor varied.

5.0  Equipment and Reagents

     5.1  Thermal Desorber.   Nu-Tech 320 or equivalent desorber  with the
          following important features:

          5.1.1  Capable of desorbing resin traps at 200-270° C.

          5.1.2  Nickel cold trap able to  be cooled to liquid nitrogen
                 temperature (-196°) and then rapidly heated to  150-250° C.
                 The upper temperature limit on the heated trap  should
                 be reached in less than 2 minutes.

          5.1.3  Heated transfer line between nickel trap and GC oven.

     5.2  Gas Chromatograph.   Varian 3700  or equivalent equipped with
          linear temperature programmer, cryogenic cooling (liquid car-
          bon dioxide or liquid nitrogen), and capillary column  capa-
          bility.

     5.3  Capillary Column.   15M fused silica, DB-5 thick film (1 micron)
          column.  Other capillary columns yielding the desired  chro-
          matographic separations may be used.

     5.4  Packed Column (optional).   6' x  2 mm I.D.  glass column packed
          with 60/80 mesh Carbo-pak C coated with 1% SP1000.   Condition
          overnight at 220° C with 20 m£/min flow rate.   Other packed col-
          umns may be used if chromatographic  separe-ion is satisfactory
          for the compounds of interest.

     5.5  Mass Spectrometer capable of scanning from 35-350 a.m.u.  in 1 sec-
          ond or less and with open-split  or direct interface for capillary.

     5.6  Data System.   Finnigan INCOS or  equivalent capable of  acquir-
          ing and storing continuous repetitive mass spectra from the
          mass spectrometer.   The system must  be  able to match unknown
          spectra to the  EPA/NIH/MSDC mass spectral  library and  integrate

-------
                                                                            C-4
          ions for quantisation.  Automated processing of the data  is
          desirable.
     5.7  Culture Tubes.  Pyrex glass screw cap tubes 25 mm x 150 mm.  Pyrex
          9825 or equivalent washed, dried, baked and fitted with Teflon-
          backed butyl rubber septa as described in Reference 12.10.
     5.8  Pyrex glass wool.  Prepared as in Reference 12.10.
     5.9  Calcium sulfate or sodium sulfate.   Anhydrous, non-indicating
          Baked at 220° C for at least 1 hour prior to use.
    5.10  5/8" Teflon Rod with V drilled hole.   Sized to hold Resin traps
          securely against septa of culture tubes.
    5.11  Resin traps as described in Reference  12.10.
6.0  Instrument Conditions
     6.1  Desorber
          6.1.1  Block Temperature 220° C - 270° C.   220° is the usual  op-
                 erating temperature.
          6.1.2  Desorber flow rate 15 m£/min helium.
          6.1.3  Cold trap temperature -190°  C (Reads  approximately 160 on
                 Nu-tech Model  320 thermal  desorber).
          6.1.4  Cryogenic trap desorb temperature  180°  C.
          6.1.5  Transfer line  temperature  180°  C.
     6.2  Gas  Chromatograph  (fused silica capillary  column).
          6.2.1  Carrier (helium)  pressure  14 psig.
          6.2.2  Initial  temperature  -20° C.
          6.2.3  Initial  hold time  2  minutes.
          6.2.4  Program  rate 5° C/min.
          6.2.5  Final  temperature  220° C.
          6.2.6  Final  hold  time 15 min.
          6.2.7  GC/MS  separator oven  240°  C.
          6.2.8  Make-up  gas flow  (for open split) 30 m£/min.

-------
                                                                       C-5
6.3  Gas Chromatograph  (packed column)
     6.3.1  Carrier  (helium) flow rate 30 mA/min.
     6.3.2  Initial  temperature 60°.
     6.3.3  Initial  hold time 4 minutes.
     6.3.4  Program  rate 8° C/minute.
     6.3.5  Final temperature 220° C.
     6.3.6  Final hold time 15 minutes.
     6.3.7  GC/MS separator oven 240° C.
6.4  Mass Spectrometer
     6.4.1  Source temperature 220° C.
     6.4.2  Mass Range 35 to 350.   Other higher mass ranges may be
            used.
     6.4.3  Scan time 0.95 seconds up, 0.05 seconds hold at the
            bottom of the scan.
     6.4.4  Electron energy 70 eV.
     6.4.5  Emission current 1.5 mA.
     6.4.6  Line-of-sight inlet 230° C.
6.5  Sample Introduction Timing Sequence
     6.5.1  Before sample introduction the desorber valve is in
            the desorb mode,  the cold trap is at liquid nitrogen
            temperature and the sample desorbtion chamber is at
            operating temperature.   Time zero is the time at which
            the GC temperature  program is started.
     6.5.2   t  = -8 min  0 sec       Insert Resin trap into thermal
                                   desorber.
     6.5.3   t  = -0 min 15 sec       Turn on Ionizer.
     6.5.4   t  =  0 min  0 sec       Start GC  oven program.   Begin
                                   mass  spectral data acquisition.
                                   Remove liquid nitrogen bath from
                                   the cold  trap.
     6.5.5   t  =  0 min 30 sec       Begin  heating nickel  cold trap
                                   and switch to inject mode on
                                   desorber.

-------
                                                                           C-6
          6.5.6  t =  2 min 30 sec      Turn off heat on nickel cold trap.
                                        Return desorber valve to desorb
                                        mode.

          6.5.7  t = 65 min  0 sec      Analysis complete.

7.0  Procedure

     7.1  At least 16 hours before analysis of a sample, traps should be
          dried.  In an organic vapor free area, transfer Tenax resin
          traps to cool, clean culture tubes containing approximately 10 g
          anhydrous sodium sulfate or calcium sulfate.   The desiccant should
          be held in place with clean glass wool.   Securely cap the culture
          tube.  This removes water adsorbed onto the Tenax during sampling.
          The culture tubes should be stored in a desiccator with activated
          charcoal adsorbent at room temperature.   This step may be omitted
          if the humidity was less than 20% during sampling.  Other drying
          techniques may be necessary with different resins which adsorb
          water more strongly.

     7.2  Set up instrument conditions as described in Section 6.

     7.3  Spike the trap with surrogate standards.   See Section 9.

     7.4  Spike the trap with 20 u£ Internal Standard.   See Section 9.

     7.5  Begin analysis.   Use the procedure described in 6.5.

     7.6  After analysis is complete,  output, and evaluate  data.

8.0  Storage and Holding Times

     8.1  Samples prior to the drying  step are stored in a  dark organic-
          free area and held at -20° C or less.   (Ref.12.12).

     8.2  The samples should be analyzed within 4 weeks of  collection.
          (Ref.12.12).

9.0  Standards

     9.1  Internal  Standard (Static)

          9.1.1  To a clean 300-m2 glass gas sampling bulb  prepurged with
                 inert gas add 9.3 u£  hexafluorobenzene and 10.3  pfi ds-
                 bromoethane.

          9.1.2  Maintain  the  bulb in  a  water bath  at 30° +0.1° C.   The
                 bulb should be in the water bath  for at least  1  hour
                 before  sampling.

          9.1.3  Withdraw  20 u£ aliquots using  a  gas-tight  syringe.   Slowly
                 inject  into the  center  of the  sample or standard  trap.
                 This injects  1000 ng  of each compound.   See calculations
                 9.1.3.1 and 9.1.3.2.

-------
                                                                       C-7
             9.1.3.1  Hexafluorobenzene has a density of 1.607 q/m£
                      or 1.607 mg/u£.
                                                        -3

                                   x onn £o X 20 US, X ——;
                                       _3
                               1.0 x 10   mg = 1000 ng
             9.1.3.2  Bromoethane has a density of 1.4606 g/m£.

                      1-4606 mg   10.3 u£ v 20 \tl x 10"3 m£ _
                         u£       300 m£  x         	ji£	
                                  -3
                          1.0 x 10   mg = 1000 ng

 9.2  Surrogate Standards (Dynamic)

      9.2.1  Prepare permeation tubes of bromochloromethane and  bromo-
             pentafluorobenzene using the procedure outlined in  Ref-
             erence  12.9.

      9.2.2  Spike surrogates onto sample and  standard  tubes immediately
             prior to analysis.   This is done  by  connecting the  Tenax
             trap to the  gas  exit flow from the permeation  chamber.
             Time to the  nearest second the length  of time  that  the trap
             has flow from  the permeation chamber going thru it.   This
             time (min) multiplied by the permeation rate (ng/min) gives
             the amount (ng)  of each  compound  on  the Tenax  trap.   The
             usual time is  3  min.

9.3   Mass  Intensity Standard (Static)

      9.3.1   A  static  standard  of  octafluorotoluene is  prepared  the same
             as the  internal  standard  (Section 9.1).  Use a  250  m£ gas-
             sampling  bulb  and  7.5 u£  octafluorotoluene.  Injecting 20 u£
             of the  gas from  the bulb  places 1000 ng of  octafluorotolu-
             ene on  the trap.   See calculation 9.3.2.

      9.3.2  Octafluorotoluene  has a density of 1.663 mg/(j£
                                 .3
                         1.0 x 10   mg = 1000 ng
9.4  Quantisation Standards (Dynamic)

     9.4.1  Prepare permeation tubes of compounds to be quantified
            as in Sec.  9.2.1.

-------
                                                                           C-8
          9.4.2  Spike standards onto blank resin traps for analysis using
                 the procedure described in 9.2.2.  Add surrogate stand-
                 ard and internal standard before beginning analysis.

          9.4.3  The amount of material spiked onto the traps is controlled
                 by how long the outflow from the permeation tubes is al-
                 lowed to flow through the resin trap.  This is a linear,
                 reproducible relationship to times at least down to I min.
                 Figure 1 shows the linearity typically achieved between
                 FID area response and sampling time.

10.0 Quantification

    10.1  Chemicals identified from their mass spectra may be quantified
          by comparison of the responses of the unknowns to the responses
          of known amounts of pure standards.   The preferred method is the
          use of relative responses and internal standards.

    10.2  Calibration is performed by analyzing a mixture of chemicals at
          known concentrations containing an internal  standard (hexafluoro-
          benzene for example) added at a fixed concentration.   The instru-
          ment responses for selected ions are measured and compared for
          each component.   A response factor is calculated for each com-
          ponent by:

          Resp.  fact.  =  Area x Ref.  Amt/(Ref.  area x  amt.)     (Eq.  1).
          Where:   (Resp.  fact.  = response factor)

                    Area      = area of ion in component
                    Ref.  Area = area of ion in internal standard
                    Ref.  Amt.  = amount of internal standard added
                    Amt.       = amount of component

    10.3  Quantification of identified chemicals is done by determining
          the areas of the appropriate ions and calculating the amount
          from equation 1 re-arranged:

               Amt = Area x Ref.  amt/(Ref.  area x amt.)         (Eq.  2).

11.0 Quality Control

    11.1  A laboratory blank spiked with the surrogate and internal
          standards is run daily before the analysis of samples.   If
          there  is a  response change between the surrogate and  internal
          standard compounds the cause is investigated and corrected
          before  analysis  of samples.   There should not be peaks in the
          blank which might interfere in the analysis.   Tenax  traps are
          easily  contaminated by solvents and  other volatile organics.

    11.2  Octafluorotoluene is  run daily before the analysis of samples
          as a check  of mass-intensity calibration.  (May be combined with
          laboratory  blank.)  Mass-intensity criteria  for 1000  ng of octa-
          fluorotoluene are given on page C-10.

-------
Absolute
Area
(GC/FID)
Response
                      i

               EFfl-HEIC
                                                L in v a r-  1 e ar t  5 q u a r e =  fit
             53253
                                                                _  IT
5140.59
                                      coefficient of determination  IP  2':  .?
      3        4        5       t        ?

Time  permeation tube outflow  is sampled  (min).
                   •?      10

                   nEHVEP, CULOPHDO
                                Figure 1.  Typical  linearity curve (area  vs time)
                                                                                                            o
                                                                                                            10

-------
                                                                        C-10
                 ID/I.                      % Relative  Abundance

                  69                              30-60
                  79                               5-15
                  93                              10-30
                 117                              40-65
                 167                              10-25
                 186                              55-85
                 217                              100
                 236                              60-85

 11.3   Calibration of  response  factors  is  done  daily  at a mid-range
       concentration.   Linearity  is determined  at  least once during
       a  set of analyses.

 11.4   The  response of the surrogate standards  is  monitored relative
       to the  internal  standard.  Any significant  deviation is inves-
       tigated and proper corrective action  is  taken  before other
       samples are run.  It is  very important to monitor this and
       correct problems immediately, since  it is not  possible to re-
       run  a sample.

 11.5   Air  sample  traps are easily contaminated.   One sample trap per
       shipping container taken to the  field will  be  tagged in the
       field as a  field blank and returned with the samples to the
       laboratory  for  analysis.

 11.6   Field spikes are air sample traps which have been spiked in
       the  laboratory, taken to the field, tagged  and returned to
       the  laboratory.  These spikes indicate sample deterioration
       due  to shipping, handling and storage.  Three field spikes
       or a number  equal to 10% of the field sampling points, which-
       ever is greater, will  be analyzed.

 11.7   Breakthrough spikes are air sample traps which have been spiked
       in the laboratory and  subsequently sampled over in the field
      with a clean sample trap as the back-up in the tandem sampling
      arrangement  [see Figure 2].  This spike indicates whether break-
      through of spiked compounds has occurred under field conditions.
      The breakthrough spike  is done  in duplicate at the upwind sampling
      point.

11.8  Table 1 lists typical  response  factors,  percent relative stand-
      ard deviations,  and lower limits  of detection.

11.9  Figure 3 is a typical chromatogram obtained using the 15 M DBS
      capillary  column.

-------
         FLOW
' _ .'i ir
/
i
[ i
vj 	 1
' 	 1 ' 	 V >*£"• '..••:'-,. . , •'•{! i
H h- irKi: ':.•'•••<.' -W f i
J_M ' m -, ;•,••:...:••• ; ;•:; ji j
""^ -1 — •-^Xv^ — ..., , ''•'": \ U_ i
i i i
\ '• r~
% Teflon Union Air Sample Trao 'v~ 	 '•
!
!
•
1 ^
_ , »
I
j
i
: J
\
*'"*,-: ' * v\
• ' : '. * ' " - .-
'• • x " ' ',

CDT l/cn n •. . TT._
5/8" Brass Union with
    Teflon ferrules
                                 Sample
Figure 2.   Breakthrough Spike
                                                                    o

-------
                                                                           c-
                             Table 2
         IDENTITIES AND CONCENTRATION OF PEAKS IN FIGURE 3
Peak #
I
2
3
4
5
6
7
8
9
10
11
12
13
14
14
15
16
17
18
19
20
21
22
23
24
25
26
HC
Compound
Carbon dioxide
Tr ichl orofl uoromethane
Acetone
trans- 1,2-Dichloroethene
Dichloromethane
1,1-Dichloroethene
1 , 1-Di chl oroethane
Hexane
Hexaf 1 uorobenzene
Tetrahydrofuran
1,2-Dichloroethane
1,1,1-Tri chl oroethane
Octafluorotoluene
Benzene (coelutes)
Methylcyclopentene (coelutes)
Carbon tetrachloride
Cyclohexane
Methyl hexane
1,2-Dichloropropane
1,1, 2-Tri chl oroethane
Toluene
Tetrachloroethene
Octane
Chlorobenzene
Ethylbenzene
o-Chlorotoluene
1, 2, 4-Tri Chlorobenzene
Unidentified alkane/alkene
Amount (ng)
***
600
980
4000
2
***
2000
620
510
2500
2
***
870
270
2
***
610
2
***
1000
***
2
***
260
150
380
950
2
***
380
150
2
***
70
***
1  Carbon dixode is an artifact of the analysis.   Identification
     is based on spectra and that it is an unretained compound.
2  New permeation tube; not calibrated.

-------
                                                                          DATA:
                                                                          CALI:
1C0.0
 R!C
C4/30/82  12:36:69
SAMPLE: 3 mil ARRAY 1.  5 HIM ARRAY 2, 5Q UL  HFB
RAtiOE: G    1,1309   LABEL:  M  0, 4.€  GUAN: A  0,  1.0  BASE: U 23,
                    fv 9 V
                    D*r
AB 121.266 {(1
CC4334 i)4
                                                                                     SCAiJS
1 TO 13^

256

89
i

i
i















t
i
i

!

i






3!
t
3
! i !

i • i


1
| 1
i








~J







\Xw,
5'
ti

B
s
II fl
II
1 ii A 1
I ii
In


5

ii ii
i ii
I i
l|
i ' i
)
i
i




jt
jj . i
i
.1 i;
! i










L ^
«
I
t








L
t


1

A

£j






/O



i
•1 '
i




ij

1
ii
1
1 -
1 ,
L« t
,,






















"' 1'










1
V]









a
c_ -/ 1 0 t' a.
, 1244 '2t

1392
1 '^1

| | . l I | 	 r 	 1 	 . w
58Q 1680 1500 $a?J
8:20 16:4!li 25:63 1M£

-------
                                                                           C-14
12.0 References

    12.1  Bertsch,  Wolfgang,  Chang,  Ray C.  and Albert Zlatkis,  "The
          Determination of Organic Volatiles in Air Pollution Studies:
          Characterization of Profiles",  Journal of Chromatographlc
          Science.  Vol.  12, pp 175-182, April  1974.

    12.2  Pellizzari,  Edo D., "Development  of Method for Carcinogenic
          Vapor Analysis in Ambient Atmospheres",  EPA-650/2-74-121, July
          1974.

    12.3  Pellizzari,  Edo D., Bunch, John E.,  and  Ben H.  Carpenter,
          "Collection  and Analysis of Trace Organic Vapor Pollutants in
          Ambient Atmospheres:  Technique for Evaluating Concentration
          of Vapors by Sorbent Media", Environmental  Science and Tech-
          nology, Vol  9, pp 552-553, 1975.

    12.4  Ibid,  pp  556-560.

    12.5  Pellizzari,  Edo D., "Development  of Analytical  Techniques for
          Measuring Ambient Atmospheric Carcinogenic Vapors",  EPA 600/2-
          75-076, November 1975.

    12.6  Pellizzari,  Edo D., "The Measurement of  Carcinogenic  Vapors in
          Ambient Atmospheres",  EPA  600/7-77-055,  June 1977.

    12.7  Pellizzari,  Edo D., "Analysis of  Organic Air Pollutants by Gas
          Chromatography and  Mass  Spectroscopy:  Final Report",  EPA
          600/2-79-057,  March 1979.

    12.8  Pellizzari,  Edo D., "Ambient Air  Carcinogenic Vapors:   Improved
          Sampling  and Analytical.  Techniques  and Field Studies",  EPA
          60012-79-081,  May 1979.

    12.9  "Volatiie Organic Air Pollutant Analysis -  Permeation Tube
          Preparation  and Calibration", NEIC,  July 1982.

   12.10  "Volatile Organic Air PoJ-Zutant Analysis -  Tenax Sample Trap Pre-
          paration  and Screening", NEIC,  March 1983.

   12.11  "Vo.Zati.Ze Organic Air PoJJutant Analysis -  Sample  Collection",
          NEIC,  March  1983.

   12.12  Pellizzari,  Edo D. ,  "AnaJyticaJ Protocol:   Personal Monitoring
          of Vapor  Phase Organic Compounds  in  Ambient Air (RTI)".

-------
PERMEATION TUBE PREPARATION AND CALIBRATION

-------
                   Volatile Organic Air Pollutant Analysis
                 Permeation Tube Preoaration  and Calibration
                              NEIC  July 1982

 1.0  Introduction

      1.1   Primary standards are necessary  to quantitatively analyze or-
           ganic  air pollutants.   Standards are  prepared  by loading sam-
           ling traps with  known amounts of chemicals  from permeation
           tubes.  This is accomplished  by passing the  effluent gas  stream
           from a  chamber containing  calibrated  permeation tubes  onto
           sampling trans identical  to  those  used in the  field.

      1.2   Permeation tubes  are  generally Teflon tubes  containing a pure
           chemical,  plugged to  form  gas tight seals at each end    The
           organic chemical  then  oermeates through  the  Teflon tubing at  a
           rate dependent upon the  temoerature and  length  of the  tube
           The  rates  are also dependent  upon  the chemical  and vary  over
           serveral  orders  of magnitude.  The permeation  rate is  deter-
           mined gravametrically.

2.0   Safety

      2.1   Many of the compounds  of interest  in  air analysis  are  toxic
           and/or  carcinogenic.   They are also volatile which  increases
           the  potential for exposure to  the compounds.  Persons  prepar-
           ing  permeation tube standards must be  aware  of  the  hazards of
           the  individual compounds handled, and  use appropriate  safety
           precautions.  All permeation  tubes should be prepared  in  a
           hood, in extreme cases other  precautions may be necessary.

     2.2  The  permeation tubes slowly emit the standard compounds.   The
          erriuent from the permeation  tubes  should be routed through a
          charcoa  trap and into a fume hood.  When weighing  the tubes,
          the analyst should keep handling  to a  minimum and avoid breath-
        -  ing fumes from the tubes.

3.0  Tube  Materials
                                                   and propyiene
                     03blwail  Tetraf1uoroethylene Po1ymer tubing 1/4"
     3.3  Teflon  Rod.   Del-F rod  3/16"  o.d.


     3-4
     3.5   Crimp  tool.   Nicopress  31-CJ  tool  to  crimp  to  1/4"  o.d.

-------
                                                                          D-2

  4.0  Permeation Chamoer



                            \ Recirculatin9 heating/cooling bath  capable
                          a temperature of 30 ± 0.1  deg.  C?      ^°u\e


       4.2   Water Jacket.  Glass water jacketed tube  with  Teflon screw  in

            Plugs  at ends.  Typical dimensions 3 cm i.d. x 20 cm
       4'3  dlTvPr? t6lA ,nta1?l6SS Steel  CflPi11ary  ^be capable of
           delivering 40-60 cc/min of N2  from  a  30 psig supply.


       4.4  Switching valve.  Teflon 2-way  solenoid valve.




                        coa                             with


 5.0  Tube Preparation
                                              Of the               .
                                          carcin°9enic,  toxic or Haz-
                           C" FE" P'U9  in  the °"e" e"" •"« erl^. a band


     5.5  Visually inspect the tube  for signs of leaking.



     5'6  COTs?aSet™Se ''J the,Derm«"<'n chamber and maintain at a
                        n                 ..               2
6.0  Tube Calibration
          ml/min1!!  ^L^f" at C°,nStant temP^ature with about 40-60
                                r -a
                    of the balance calibration  and/or  repair



-------
                                                                            D-3
      6.5  Monitor the weight changes of calibrated tubes every 4-6 weeks
           for the life of the tube.

 7.0  Calculations

      7.1  Average Rate

           7.1.1   Rate:

                  Rate =     weight change   (ng)
                         minutes  between weighings


           7.1.2   Typically weight  change is  less than  10 mg  and  time
                  between weighings is  20000-40000  minutes (2-4 weeks).

           7.1.3   Average the last  5  stable rates.

      7.2  Regression  Rate


           7.2.1   Tabulate the weight of the  tube vs. time from the  point
                  the  rate stabilized,  or for  the last  20  stable weighings.

           7.2.2   Using least squares techniques, calculate the slope of
                  the  weight  vs.  time data.   The slope  is  the permeation
                  rate.   Also calculate  the correlation coefficient  as an
                  indication  of the stability of the calibration data.

      7.3   Percent Relative Standard Deviation (%RSD)

           7.3.1  SRSD  =  Standard Deviation
                              Average      x  IUU/J

           7.3.2  The percent relative standard deviation  should be  10%
                 or less  for stable tubes.

8.0  Other Permeation Devices
                              -u  Pfrmeation ^vices may be used such
                     ,  permeation bags  or drilled rod .

9.0  References


                                       0n'  C'  Ana1.ytical  Chemistry. 39,


                '  P" Wood>  R->  Chemistry  and  Industry.  Dec.  28,  1968,


     3.    Scaringelli   Frank  P.,  O'Keefe, Andrew  E. ,  Rosenberg,  Ethan
          Be11'  John p-»  Analytical  Chemistry.  Vol. 42, 871  (1970)

     4.    Analytical Chemistry. 49,  1278 (1977).


                                      vapors  '

-------
                                                                               D-4
                                    TABLE I
  Compound
  Acetone
  Acrylonitrile
  Benzene
  Bis(2-chloroethyl)ether
  Bromochloromethane
  Bromoethane-dr
  Bromoform
  Bromopentaf1uorobenzene
  Carbon  tetrachloride
  Chlorobenzene
  Chloroform
  1,4-Dichlorobutane
  1,1-Dichloroethane
  1,2-Dichloroethane
 1,1-Dichloroethene
 1,2-trans-Dichloroethene
 Dichloromethane
 1,2-Dichloropropane
 Ethyl benzene
 Hexane
 Tetrachloroethene
 Toluene
 1,2,4-TriChlorobenzene
 1,1,1-Trichloroethane
 1,1,2-Trichloroethane
 Trichloroethene
 Trichlorofluoromethane
Cas ^
67-64-1
107-13-1
71-43-2
111-44-4
74-97-5
Not Available
75-25-2
344-04-7
56-23-5
108-90-7
67-66-3
110-56-5
75-34-3
107-06-2
75-35-4
156-60-5
75-09-2
78-87-5
100-41-4
110-54-3
127-18-1
108-88-3
120-82-1
71-55-6
79-00-5
79-01-6
75-69-4
Tube Type
TFE
TFE
TFE
TFE
FEP
TFE
TFE
FEP
TFE
TFE
TFE
TFE
TFE
TFE
FEP
FEP
TFE
TFE
TFE
TFE
TFE
TFE
TFE
TFE
• - TFE
TFE
FEP
Rate/length*
ng/min/cm
64
88
40
2
25
170
4
71
20
25
110
4
40
58
64
150
320
8
10
37
62
26
4
9
9
200
100
*This should only be used as a guide.  Each batch of tubing will be different.

-------
TENAX-G(T SAMPLE TRAP PREPARATION AND SCREENING

-------
                      ENVIRONMENTAL  PROTECTION AGENCY
                 NATIONAL  ENFORCEMENT  INVESTIGATIONS CENTER
                       Box 25227  Denver  Federal  Center
                           Denver, Colorado  80225
                         e Organic Air Pollutant Analysis
               Tenax   Sample Trap Preparation and  Screening
                                 March 1983
1.0   Introduction
      1.1  Sampling for organics  in air  is performed by drawing air
          through a glass tube packed with the porous polymer resin
          Tenax-GC.  The traps and resin must be thoroughly cleaned
          before use to minimize the trap background.  Clean traps
          ready for field use must also be carefully packed in clean
          glass tubes to avoid contamination during handling.
2.0   Materials
      2.1  Glass sampling traps.  Pyrex glass traps constructed as shown
          in Figure 1.
      2.2  Resin.  Tenax-GC, 35/60 mesh.
      2.3  Glass wool.
      2.4  Culture tubes.  Pyrex glass screw cap tubes 25 mm x 150 mm.
          Pyrex 9825-20X or equivalent.
                ®
      2.5  Teflon -backed silicone septa.  Pierce 12722 or equivalent.
      2.6  Bakelite screw caps to fit culture tubes.  Pierce 13219 or
          equivalent.
      2.7  Desiccator.  Glass desiccator with activated charcoal
          adsorbant.
      2.8  Virgin gallon paint cans with pressure fit lids.
      2.9  Polyurethane foam.  Used for packaging of culture tubes to
          avoid breakage.
      2.10 Teflon spacers made from 5/8" Teflon rod with V drilled hole.
          Sized to hold traps securely against septa of culture tube.
3.0   Resin Preparation
      3.1  Extract new and used Tenax-GC with methanol followed by pen-
          tane in a soxhlet extractor.   Extract with each solvent 18
          hours.
      3.2  Dry the resin under vacuum for a least 6 hours.
®  Registered trademark; appears hereafter without the ®.

-------
                                                                            E-2
      3.3  Sieve  the  dried  resin  to  the  35/60  mesh  particle  size  range.

      3.4  Seal the cleaned and sieved resin  in  a glass  jar  capped  with
           a  Teflon liner.   Store in a desiccator containing activated
           carbon.

 4.0   Trap  and Container  Cleanup

      4.1  Wash new and used  glass sampling traps and culture tubes with
           lab detergent  and  hot  tap water.  Rinse  at least  three times
           with organics-free water.  Rinse with methanol and let air
           dry.

      4.2  Bake the cleaned tubes in an  oven at 220° C for at least 1 hour.
           Remove from the  oven and  store  in a desiccator containing acti-
           vated carbon.

      4.3  Place glass wool in boiling water for 15 minutes.    Remove glass
           wool from boiling  water,  drain, and place on absorbent paper
           towels to remove excess water.  Rinse by dipping  in acetone and
           placing in a soxhlet extractor.  Soxhlet extract  glass wool with
           methanol at least  8 hours, air dry and bake in an  oven at 200° C
           for at least 1 hour.   Spread  the glass wool out in a baking dish
           and place in a muffle furnace at 450° C for 1 hour.  Remove from
           the oven, cool  and store  in a desiccator containing activated
           charcoal.

     4.4   Bake Teflon-backed septa  in an oven at 80° C for 30 minutes.
           Remove from the oven and  store in a desiccator containing
           activated charcoal.

     4.5   Bake paint cans in oven at 100° C for 1 hour.

     4.6   Sohxlet extract foam in methanol for 18 hours.  Drain foam and
          dry under vacuum for 24 hours.

5.0  Trap Preparation

     5.1  Pack about  a 1-cm plug  of glass wool into the  trap and weigh
          it.  Add about  7 cm of  cleaned Tenax-GC.   Lightly  tap the trap
          on the  bench to pack  the resin.   Weigh the trap and adjust the
          weight  of the Tenax to  0.8 ± 0.02 gram by adding or removing
          small  amounts of the  polymer.   Add  another 1-cm glass wool  plug
          to hold the resin in  place.

     5.2  Condition each  trap by  placing it in the  bakeout manifold.
          The manifold is an apparatus which  provides  flow to six traps
          simultaneously  and can  be  placed in  a  GC  oven.  Connect a k"
          to 1/16"  reducing filter  on the top  of each  trap and turn on
          the helium  flow.   Measure  the  flow  on  each trap to assure it
          is  20-30  m£/minute.  Set  the GC for  bakeout conditions  of
          270°  for  30 minutes.  Set  the  oven  initial  temperature  for
          cooldown  to 35°.   After cooldown, turn  off the helium flow

-------
                                                                           E-3
     5.3
           and  place  the  traps  in  culture  tubes  with  Teflon spacers.
           Cap  the  culture  tube using  a  Teflon-lined  septum [Figure 1].
           Record the bakeout batch  number on  each  tube  label.

           Tubes are  stored in  desiccators containing activated  charcoal
           until ready for  shipment.
6.0  Screening

     6.1  Screen at least one trap from each bakeout batch on a thermal
          desorber system connected to an FID-GC.   Thermal desorber con-
          ditions are the same as those used for GC/MS analysis.   Elec-
          trometer setting is 10-12 amps and attenuation 32.   Label each
          chromatogram with a GC run number and store the chromatograms
          by ascending number for future reference.   Figure 2 shows a
          typical acceptable screening chromatogram.
                 GC Conditions
                                                 Desorber Conditions
              Column:

                Phase:   DB-5 or Equivalent
                Length:   15 meter

              Injection  Temp:   110° C
              Detection  Temp:   260° C
              Oven:   Initial  Temp.  35°  C
              Initial  Time:   3  min.
              Prog.  Rate:   15 deg/min.
              Final  Tern:   230°  C
              Final  Hold:   15 min.
              Cap  Pressure:   10 PSI
                                             Temperatures:
                                                   Line: 150° C
                                                   Block:  220° C
                                                   Trap:  180° C
                                             Time: 0:00 min. Desorb at
                                                     LN2 Temp, on the
                                                     trap.
                                                   8 min. - Start GC and
                                                     Integrator and Chart
                                                     and Lower LN2 bath.
                                                   8 min-30 sec. - Valve
                                                     to Inject trap switch
                                                     to heat.
                                                  10 min-30 sec. - Valve
                                                     to desorb trap switch
                                                     to cool.

     6.2  Mark the label on the culture tube with the GC run number of
          the screening chromatogram.   If the trap is clean, the batch
          is acceptable for use.

7.0  Packaging for Shipment
    7.1
          Place culture tubes containing Tenax traps in special foam inserts
          in 1-gallon paint cans.   Insert a charcoal-filled pouch around the
          top of the foam insert and pour a few drops of liquid nitrogen
          onto the foam.   After a few seconds place the lid securely on the
          paint can.   If excessive bulging of the can occurs, immediately
          remove the lid to release the nitrogen gas.   Secure can lids
          with clips.   Place a label on the can lid indicating the number
          and types of traps inside.  Indicate the type of adsorbant and
          whether spikes or blanks are present.

-------
Attachment to:   Sample Trap Preparation
                                                                          E-4
             &  ^
           4
          I 5 fr)m
        _£_.
^ /A
                                                 JvrV_
               Figure  1.  Sampling trap and culture  tube holder design.

-------
         Figure 2.     Typical FID screening of a Tenax trap after bakeout at 270°C.   This  result  is
                       acceptable for sample traps to be used for GC/MS analysis.   Because  of the  dif-
                       ference in sensitivity between FID and MS, only the C02 peak would appear in an
                       MS analyis.  Chromatographic conditions are given in Sec.  6.1.
                       Ill


                              -ll-rn+H-r;
                                I !  '  ; i : ;   '
                      ._].„.
                       I I '
                                                                            ~rt"TTTJT
                                                I

: !|"i"! ii'M ! J;'M '! i" il!"::"|
. : i .:•:.!. . . ;
. 1 , i ' , • •
1 i : • i
	 I f • , • 1 : • •
: : ; i i ! : i ! : • : , i ; - !
	 L. ...„! 	 - ... : .
; i ' |
'. I ' i
1
r. • ;

i
1
; '• '

                                                                                  I i I :
                                                                                                           — rv-i itpi'-r-'
                                                                                                      ^^.__3m^I I I   ' *• (
                                    I
I	!_..  _j._. ._L.L.LLill
H-
-H
                                i    i
         i  i '
1 • . ; . ' '.

• i :
1 - i '
1' : • i 	
1 i I i
MIL

'


i : .
> J,
! . ir
i : I '
I : : i ! : . : i
1
!-..,.,. . . ...
'
     -L^LL
       , i   i

                                                                                                                 -'- f—	
nii«-3-
                                                          LlP-iFAR
                                                             !  H-l
                                                        I
                                                   !     I

                                                   : :    i  •
                                                   _l_ _  	I
                                                       "i r  —
                                                   j  : i ; |  i
                                                       -1 •'
                                                                                                              3

-------
                                                                           E-6
8.0  References

     8.1  "Selection and Evaluation of Sorbant Resins for the Collection
          of Organic Compounds",  EPA-600/7-77044,  April  1977.

     8.2  "Development of Method  for Carcinogenic  Vapor  Analysis  in Am-
          bient Atmospheres",  EPA-650/2-74-121,  July 1974.

-------