Test Methods for Evaluating Solid Waste
      Physical/Chemical Methods
 Final (Promulgated) Updates II and MA
              Volume 2

-------
                                 METHOD 8000A

                              GAS CHROMATOGRAPHY
1.0   SCOPE AND APPLICATION
      1.1   Gas  chromatography is  a quantitative  technique  useful  for  the
analysis  of organic  compounds  capable of  being  volatilized without  being
decomposed or  chemically  rearranged.   Gas chromatography  (GC),  also  known as
vapor phase chromatography (VPC), has two subcategories distinguished by: gas-
solid chromatography  (GSC),  and  gas-liquid  chromatography (GLC)  or gas-liquid
partition chromatography  (GLPC).   This  last group  is  the most commonly used,
distinguished by type of column adsorbent or packing.

      1.2   The chromatographic methods are recommended for use only by, or under
the close supervision of, experienced residue analysts.


2.0   SUMMARY OF METHOD

      2.1   Each organic analytical method that follows provides a recommended
technique  for  extraction,  cleanup,  and  occasionally,  derivatization  of  the
samples to be analyzed.  Before the prepared sample is introduced into the GC,
a procedure for standardization must be followed to determine the recovery and
the/ limits of  detection  for the  analytes  of  interest.    Following  sample
introduction into the GC, analysis proceeds with a comparison of sample values
with standard values.  Quantitative analysis  is achieved through integration of
peak area or measurement of peak height.


3,0   INTERFERENCES

      3.1   Contamination by carryover can occur whenever high-concentration and
low-concentration samples are  sequentially analyzed.  To  reduce carryover, the
sample  syringe or purging device must be rinsed out between samples with water
or solvent.  Whenever an unusually concentrated sample is encountered, it should
be r followed by  an  analysis  of a solvent blank or of  water to check for cross
cb'ritami nation.  For volatile  samples containing large amounts of water-soluble
materials,  suspended  solids, high  boiling  compounds or high  organohalide
concentrations,  it may be necessary  to  wash out the syringe or purging device
with a  detergent solution, rinse it with distilled water,  and then dry  it  in  a
     2 oven between analyses.
4,0   APPARATUS AND MATERIALS

      4.1   Gas chromatograph - Analytical system complete with gas chromatograph
suitable  for  on-column  injections  and  all  required  accessories,   including
.detectors, column  supplies,  recorder,  gases,  and  syringes.   A data system for
measuring peak height and/or peak areas  is recommended.
      4.2
Other packed or capi
Gas chromatographic columns - See the specific determinative method.
 or capillary (open-tubular) columns may be  used if the requirements

                      8000A  -  1                         Revision 1
                                                         July 1992

-------
of Section 8.6 are met.


5.0   REAGENTS

      5.1   See the specific determinative method for the reagents needed.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See the  introductory material  to this  chapter,  Organic Analytes,
Section 4.1.


7.0   PROCEDURE

      7.1   Extraction - Adhere to those procedures specified in the referring
determinative method.

      7.2   Cleanup and separation - Adhere to those procedures specified in the
referring determinative method.

      7,3   The recommended gas chromatographic columns and operating conditions
for the instrument are specified  in the referring determinative method.

      7.4   Calibration

            7.4.1 Establish gas chromatographic  operating  parameters equivalent
      to those indicated in Section 7.0 of the determinative method  of interest.
      Prepare   calibration  standards  using   the   procedures  indicated  in
      Section 5.0  of  the  determinative method of  interest.    Calibrate the
      chromatographic  system  using   either  the  external standard  technique
      (Section 7.4.2)  or the internal standard  technique  (Section  7.4.3).

            7.4.2 External  standard calibration procedure

                  7.4.2.1      For each analyte of interest, prepare calibration
            standards  at a minimum of five concentrations by adding volumes of
            one or more  stock  standards  to  a volumetric flask and diluting to
            volume with an appropriate solvent.   One of the external standards
            should be  at a concentration near,  but  above, the method detection
            limit.  The other  concentrations should correspond to  the expected
            range of concentrations found in real samples or should define the
            working range  of the  detector.

                  7.4.2.2      Inject  each   calibration  standard  using  the
            technique that will be used to introduce the actual  samples  into the
            gas chromatograph  (e.g.  2-5  /xL injections, purge-and-trap, etc.).
            Tabulate peak  height or  area responses against  the mass injected.
            The results  can be  used  to  prepare a  calibration  curve  for each
            analyte.   Alternatively,  for samples that are introduced into the
            gas chromatograph  using a syringe, the ratio of the response to the
            amount  injected,  defined as  the calibration factor  (CF),  can be
            calculated for each analyte at each  standard concentration.  If the

                                   8000A -  2                        Revision  1
                                                                     July 1992

-------
percent relative standard deviation (%RSD) of the calibration factor
is less  than 20%  over  the working  range,   linearity through  the
origin can be  assumed,  and the average  calibration  factor  can be
used in place of a calibration curve.

      C.Hbr.t1« factor                °  *
                                                ograms)

      *     For multi response pesticides/PCBs, use the total area of
            all peaks used for quantitation.

      7.4.2.3     The  working  calibration  curve  or  calibration
factor must be verified on each  working day  by the injection of one
or more  calibration  standards.   The frequency  of  verification is
dependent on the detector. Detectors, such  as the electron capture
detector,  that  operate   in  the   sub-nanogram  range  are  more
susceptible to changes in  detector  response  caused by GC column and
sample   effects.     Therefore,   more  frequent  verification  of
calibration is  necessary.   The  flame ionization detector  is  much
less  sensitive  and requires  less  frequent  verification.    If the
response for any analyte varies  from the predicted response by more
than  +  15%,  a  new calibration  curve  must be  prepared for  that
analyte.   For methods 8010,   8020,  and 8030,  see Table 3  in  each
method for calibration and quality  control  acceptance criteria.
      Percent Difference = -  x 100
                             *i

where:

      R,    =     Calibration Factor from first analysis.

      R2    =     Calibration Factor from succeeding analyses.

7.4.3 Internal standard calibration procedure

      7.4.3.1     To use this approach,  the analyst must select one
or more internal standards that are similar in analytical behavior
to the compounds of interest.  The analyst must further demonstrate
that  the  measurement  of the internal standard is  not  affected by
method  or matrix  interferences.    Due   to  these  limitations,  no
internal standard applicable to all samples can be suggested.

      7.4.3.2     Prepare calibration standards at a minimum of five
concentrations for each analyte of interest by adding volumes of one
or more stock standards to a volumetric  flask.  To each calibration
standard,  add a  known constant  amount of  one  or more  internal
standards and dilute to volume with  an appropriate  solvent.  One of
the  standards should  be  at  a concentration  near, but  above,  the
method detection limit.  The other concentrations should correspond
to the  expected  range of concentrations found in  real  samples or
should define the working range of the detector.
                       8000A -  3                         Revision 1
                                                         July 1992

-------
            7.4.3.3     Inject each calibration  standard  using the same
      Introduction technique that will  be  applied  to the  actual samples
      (e.g. 2 to  5 pi Injection, purge-and-trap, etc.). Tabulate the peak
      height or area responses against  the  concentration of each compound
      and  internal  standard.   Calculate response  factors  (RF)  for each
      compound as follows:

            RF    «     (AsCis)/(AisCs)

      where:

            As    »     Response for the analyte to be measured.

            Ajs    »     Response for the internal standard.

            CJ8    *     Concentration of the internal standard, M9/L.

            C.    »     Concentration  of  the  analyte  to be  measured,
                        M9/L.

            If the  RF  value over the  working  range  is constant  (<  20%
      RSD), the RF can  be  assumed  to be invariant,  and the average RF can
      be used for calculations.  Alternatively,  the  results can be used to
      plot a calibration curve of response ratios,  As/Ais versus RF.

            7.4.3.4     The  working  calibration curve  or  RF  must  be
      verified on  each working day  by the  measurement  of  one or more
      calibration standards.  The frequency of verification  is dependent
      on the detector.   Detectors, such as  the electron capture detector,
      that  operate  in  the  sub-nanogram range  are more  susceptible to
      changes in  detector  response caused by GC  column and sample effects.
      Therefore,  more frequent verification of calibration is necessary.
      The  flame  ionization  detector is much  less sensitive and requires
      less frequent verification.   If the response  for any analyte varies
      from the predicted  response by more  than  ± 15%, a new calibration
      curve must be prepared for that compound.  For methods 8010, 8020,
      and  8030,  see  Table 3 in each method  for  calibration  and quality
      control acceptance criteria.

7.5   Retention time windows

      7.5.1 Before establishing windows, make sure the GC system  is within
optimum  operating  conditions.'   Make   three  injections  of  all  single
component  standard  mixtures  and multiresponse  products   (i.e.  PCBs)
throughout the course  of  a  72  hour period.  Serial injections  over less
than a 72 hour period result in retention time windows that are too tight.

      7.5.2 Calculate the standard deviation  of  the three retention times
(use any function of retention time; including absolute retention time, or
relative  retention  time)  for  each  single component  standard.    For
multiresponse  products,  choose  one  major peak from the envelope  and
calculate  the  standard deviation  of the three  retention  times for that
peak.    The  peak  chosen  should  be   fairly  immune  to  losses  due  to
degradation and weathering  in samples.

                             8000A -  4                         Revision  1
                                                                July 1992

-------
            7.5.2.1     Plus or minus three times the standard deviation
      of the  retention times for each standard will  be used to define the
      retention time window; however, the experience of the analyst should
      weigh  heavily  in  the  interpretation  of  chromatograms.    For
      multiresponse analytes  (i.e.  PCBs),  the  analyst  should  use  the
      retention  time  window,   but  should  primarily  rely  on  pattern
      recognition.

            7.5.2.2     In those cases where the  standard deviation for a
      particular  standard  is  zero,  the  laboratory must  substitute  the
      standard deviation of a close eluting,  similar compound to develop
      a valid retention  time window.

      7.5.3 The laboratory must  calculate retention  time windows for each
standard on each GC column and whenever a new GC column is  installed.  The
data must be retained  by the laboratory.

7.6   Gas chromatographic analysis

      7.5.1 Introduction of organic compounds  into the  gas chromatograph
varies depending on the volatility of the compound.   Volatile organics are
primarily introduced by  purge-and-trap (Method  5030).  However, there are
limited  applications  (in  Method   5030)  where  direct   injection  is
acceptable.   Use of Method 3810 or 3820  as  a  screening  technique  for
volatile organic  analysis  may  be valuable  with  some sample  matrices to
prevent  overloading and  contamination of the  GC  systems.   Semi volatile
organics are introduced  by direct injection.

      7.6.2 The appropriate detector(s)  is  given in the specific method.

      7.6.3 Samples are  analyzed  in a  set referred  to as  an analysis
sequence.   The  sequence  begins  with instrument  calibration  followed by
sample   extracts   interspersed   with  multi-concentration   calibration
standards.  The sequence  ends when the set of samples has been injected or
when qualitative and/or quantitative QC  criteria are exceeded.

      7.6.4 Direct Injection -  Inject 2-5 /nL  of the sample extract using
the solvent flush technique, if the extract  is manually injected.  Smaller
volumes  (1.0 /uL) can be  injected, and the solvent flush technique is not
required, if automatic devices  are employed.   Record the volume injected
to the nearest 0.05 pi and the resulting peak  size  in area units or peak
height.

      7.6.5 If the responses exceed the linear range of the system, dilute
the extract and reanalyze.   It is  recommended that extracts be diluted so
that all  peaks  are on scale.   Overlapping peaks  are not  always evident
when  peaks are  off  scale.    Computer   reproduction  of  chromatograms,
manipulated to ensure all  peaks  are  on  scale  over a 100-fold range,  are
acceptable  if  linearity  is demonstrated.   Peak  height  measurements  are
recommended over peak  area integration when  overlapping peaks cause errors
in area  integration.

      7.6.6 If   peak  detection   is  prevented   by  the  presence   of
interferences, further cleanup  is required.

                            8000A -  5                         Revision 1
                                                               July 1992

-------
            7.6.7 Examples of chromatograms  for  the  compounds  of interest are
      frequently available in the referring analytical method.

            7.6.8 Calibrate  the system  immediately   prior  to  conducting  any
      analyses  (see  Section  7.4),   A mid-concentration  standard  must  also be
      injected  at  intervals specified  in the method and  at  the end  of  the
      analysis  sequence.    The  calibration  factor   for  each  analyte  to  be
      quantitated,  must not exceed a 15% difference when compared to the initial
      standard  of  the  analysis  sequence.    When  this criterion  is  exceeded,
      inspect  the  GC   system  to determine  the  cause  and perform  whatever
      maintenance  is necessary  (see Section  7.7)  before recalibrating  and
      proceeding with sample analysis.  All samples that were injected after the
      standard exceeding  the criterion  must be reinjected  to  avoid  errors in
      quantitation,  if  the  initial  analysis  indicated the  presence of  the
      specific target analytes  that exceeded the criterion.

            7.6.9 Establish daily retention time windows  for each analyte.   Use
      the retention time for each analyte  from Section  7.6.8 as the midpoint of
      the window  for that  day.   The daily  retention time window equals  the
      midpoint ± three times the standard deviation determined  in Section  7.5.

                  7.6.9.1     Tentative  identification of an analyte occurs when
            a peak from a sample extract falls withtn the daily retention  time
            window.  Normally, confirmation is required:  on a second GC column,
            by  GC/MS   if   concentration   permits,   or  by  other  recognized
            confirmation techniques.   Confirmation may not be necessary if the
            composition  of the  sample  matrix  is well  established  by  prior
            analyses.

                  7.6.9.2     Validation of GC system qualitative performance:
            Use the  mid-concentration  standards  interspersed  throughout  the
            analysis  sequence (Section  7.6.8) to  evaluate  this criterion.   If
            any of  the standards  fall  outside their daily retention time window,
            the system  is out of control.  Determine the cause of the problem
            and correct it  (see  Section 7.7).  All  samples that  were injected
            after  the  standard  exceeding the criteria must be  reinjected to
            avoid false negatives and possibly false positives.

      7.7   Suggested chromatography system maintenance - Corrective measures may
require any one or more of the  following remedial  actions.

            7.7.1  Packed columns - For  instruments with  injection port  traps,
      replace the  demister trap,  clean,  and deactivate  the glass injection  port
      insert or replace with a  cleaned  and  deactivated insert.    Inspect  the
      injection end of the column and remove  any foreign material  (broken glass
      from the rim of the  column or pieces of septa).   Replace the glass  wool
      with fresh deactivated glass wool.   Also,  it may be  necessary  to  remove
      the first few millimeters  of the packing material if any discoloration is
      noted, also  swab  out the  inside  walls  of  the  column if any residue is
      noted.  If these procedures fail to eliminate the degradation problem, it
      may be  necessary  to  deactivate  the metal  injector   body  (described  in
      Section 7.7.3)  and/or repack/replace the column.
                                  8000A  -  6                         Revision 1
                                                                     July 1992

-------
      7.7.2 Capillary columns - Clean and deactivate the glass injection
port insert or replace with a cleaned and deactivated insert.  Break off
the first few inches, up to one  foot,  of the injection  port side of the
column.   Remove  the  column  and  solvent  backflush according  to  the
manufacturer's instructions.   If these  procedures  fail  to eliminate the
degradation problem, it may be  necessary to deactivate the metal injector
body and/or replace the column.

      7.7.3 Metal  injector body  -  Turn off  the  oven  and  remove  the
analytical column when  the  oven  has  cooled.   Remove the glass injection
port insert (instruments with  off-column injection or Grob).  Lower the
injection port temperature  to  room temperature.   Inspect the injection
port and remove any noticeable foreign material.

            7.7.3.1     Place a  beaker  beneath the injector port inside
      the GC oven.  Using a wash  bottle,  serially rinse the entire inside
      of the  injector port with acetone and  then  toluene;  catching the
      rinsate in the beaker.

            7.7.3.2     Prepare a solution of deactivating agent (Sylon-CT
      or equivalent) following manufacturer's directions.  After all metal
      surfaces inside the injector body have been thoroughly coated with
      the deactivation  solution,  serially rinse the  injector body with
      toluene, methanol, acetone, and hexane.  Reassemble the  injector and
      replace the GC column.

7.8   Calculations

      7.8.1 External  standard  calibration  -  The  concentration  of each
analyte  in  the sample  may  be determined  by calculating the amount of
standard purged or injected, from the peak response, using the calibration
curve  or  the calibration  factor  determined  in  Section  7.4.2.    The
concentration of a  specific analyte is calculated  as follows:

      Aqueoussamples

      Concentration  (Mg/L)  = [(Ax)(A)(Vt)(D)]/[(As)(V,)(V,}]
where:
      Ax    =     Response for the analyte in the sample,  units may be in
                  area counts or peak height.

      A     »     Amount of standard injected or purged, ng.
      A     »     Response  for  the  external  standard,  units same as for

                  AX-

      V,    »     Volume  of extract  injected,  fiL.   For purge-and-trap
                  analysis, V( is not applicable and therefore - 1.

                  Dilution  factor,  if dilution  was made  on  the  sample
                  prior  to  analysis.   If  no  dilution was  made,  0=1,
                  dimensionless.

                             8000A - 7                        Revision  1
                                                               July  1992

-------
      Vt    -     Volume  of  total   extract,   /nL.    For  purge-and-trap
                  analysis, Vt is not applicable and therefore - 1.

      Vt    -     Volume of sample extracted or purged, ml.


Nonaoueous samples

      Concentration (Mg/kg) -  [(Ax)(A)(Vl)(D)]/[(A.)(V,)(W)]

where:

      W     -     Weight of sample extracted or purged, g.  The wet weight
                  or dry weight may be used, depending upon  the  specific
                  applications of the data.

      Ax, At,  A, Vt, D,  and  V, have the same  definition  as for aqueous
samples when a solid sample is purged (e.g., low  concentration  soil)  for
volatile  organic analysis  or  for  semi volatile  organic  and pesticide
extracts.   When the nonaqueous  sample  is  extracted  for  purge and trap
analysis, V, -  volume of methanol extract added to reagent  water for  purge
and trap analysis.

      7.8.2 Internal standard calibration - For each analyte  of interest,
the concentration of that analyte in the sample is calculated as  follows:

      Aqueous  samples

      Concentration (Mg/L) -  [(A8)(C,.)(0)]/[(Aii)(RF)(V1)]

where:

      Ax    -     Response of the analyte being measured,  units may  be in
                  area counts  or peak height.

      Cia   -     Amount of internal standard  added  to extract  or volume
                  purged, ng.

      D     -     Dilution factor,  if  a dilution was made on  the sample
                  prior  to  analysis.    If  no  dilution was made,  D =  1,
                  dimensionless.

      Ais   -     Response of  the internal standard,  units same  as  Ax.

      RF    -     Response factor  for  analyte, as determined  in Section
                  7.4.3.3.

      •Vt    -     Volume of water extracted or purged, ml.

      Nonaaueous samples

      Concentration (Mg/kg)  -  [(As)(Cis)(D)]/[(Ais)(RF)(Ws)]
                             8000A - 8                         Revision 1
                                                                July 1992

-------
      where:

            Ws    -     Weight of sample extracted, g.  Either a dry weight or
                        wet weight  may be  used,  depending upon  the  specific
                        application of the data.

            AS»  Cis,  D, Ajs,  and  RF  have the  same definition as  for  aqueous
      samples.


8.0   QUALITY CONTROL

      8.1   Each laboratory that  uses  these methods  is required  to operate a
formal quality control  program.  The minimum requirements of this program consist
of an initial  demonstration of laboratory capability and an ongoing analysis of
spiked samples  to  evaluate and document quality data.  The  laboratory should
maintain records to document the  quality of the data  generated.   Ongoing data
quality checks are compared with established performance criteria to determine
if the results of analyses meet the performance characteristics  of the  method.
When results  of sample spikes  indicate  atypical method  performance,  a  quality
control check standard should  be analyzed to  confirm that the measurements were
performed in an in-control mode of operation.

      8.2   Before  processing  any  samples,  the  analyst   should  demonstrate,
through the analysis of a  reagent  blank, that interferences from the analytical
system, glassware, and reagents are under control.   Each time a  set of  samples
is extracted  or. there  is  a change in reagents, an organic-free  reagent water
blank  should   be   processed   as  a  safeguard  against   chronic  laboratory
contamination.  The blank  samples should be carried through  all  stages of the
sample preparation and measurement steps.

      8.3   For each analytical batch (up to  20 samples),  a  reagent  blank, matrix
spike, and duplicate or matrix spike duplicate should be analyzed (the frequency
of the spikes  may be different  for  different monitoring programs).   The blank and
spiked samples should be carried through all stages of the sample preparation and
measurement steps.

      8.4   The experience of  the  analyst  performing  gas chromatography  is
invaluable to the  success  of the methods.  Each day that analysis  is performed,
the  daily  calibration  sample  should  be  evaluated  to  determine  if  the
chromatographic system is  operating properly.   Questions  that should  be asked
are: Do  the peaks  look normal?;  Is the response  obtained comparable  to  the
response  from previous calibrations?    Careful  examination  of   the  standard
chromatogram  can  indicate whether the  column  is  still  good, the injector  is
leaking, the  injector  septum needs replacing, etc.   If any changes are  made to
the system (e.g. column changed), recalibration of  the  system should take place.

      8.5   Required instrument QC

            8.5.1  Step 7.4 requires that the %RSD vary by < 20% when comparing
      calibration  factors  to  determine  if  a five point  calibration curve  is
      linear.
                                   8000A  -  9                         Revision 1
                                                                     July 1992

-------
            8.5.2 Section 7.4 sets a limit of ±15% difference  when comparing
      daily response  of a  given analyte  versus  the  initial  response.   For
      Methods 8010, 8020, and 8030,  follow the guidance on limits specified in
      Section 7.4.3.4.   If the limit is exceeded,  a new  standard curve should be
      prepared unless  instrument maintenance  corrects the  problem for  that
      particular analyte.

            8,5.3 Step 7.5 requires the establishment of retention time windows.

            8.5.4 Section 7.6.8 sets  a  limit of ± 15% difference when comparing
      the response from the  continuing calibration  standard of a given analyte
      versus any succeeding  standards analyzed during an  analysis sequence.

            8.5.5 Step  7,6.9.2  requires  that all  succeeding standards in an
      analysis sequence  should  fall within  the daily retention  time  window
      established by the first standard of the sequence.

      8.6   To  establish the  ability  to generate acceptable  accuracy  and
precision, the analyst should perform the following  operations.

            8.6.1 A quality control   (QC) check sample  concentrate  is  required
      containing each analyte of interest.  The QC check sample concentrate may
      be  prepared  from  pure  standard materials,  or  purchased as  certified
      solutions.   If prepared by the  laboratory, the  QC check sample concentrate
      should be  made  using  stock standards prepared independently from those
      used for calibration.

                  8.6.1.1     The .concentration  of  the  QC   check   sample
            concentrate  is   highly   dependent   upon   the  analytes   being
            investigated. Therefore, refer to Method 3500, Section 8.0 for the
            required concentration of the QC  check  sample concentrate.

            8.6.2 Preparation of QC  check samples

                  8.6.2.1     Volatile organic analytes (Methods 8010, 8020, and
            8030) - The QC  check sample is prepared  by  adding 200 ptL of the QC
            check sample concentrate (Step 8.6.1)  to 100  ml of water.

                  8.6.2.2     Semivolatile organic analytes (Methods 8040, 8060,
            8070, 8080,  8090, 8100,  8110,  and 8120) - The QC check  sample is
            prepared by adding 1.0 ml of the  QC check sample concentrate (Step
            8.6.1) to each  of four 1-L aliquots of  water.

            8.6.3 Four aliquots of the well-mixed QC check sample are analyzed
      by the same procedures used to analyze actual samples  (Section 7.0 of each
      of the methods).   For volatile organics, the preparation/analysis process
      is purge-and-trap/gas  chromatography.  For semivolatile organics,  the QC
      check samples should undergo solvent extraction (see Method 3500) prior to
      chromatographic analysis.

            8.6.4 Calculate  the average recovery (x) in u.g/L, and the standard
      deviation of the recovery (s)  in  u.g/U for each analyte of interest using
      the four results.
                                  8000A - 10                        Revision 1
                                                                     July 1992

-------
            8.6.5 For  each  analyte  compare s  and x  with the  corresponding
      acceptance criteria for precision and accuracy, respectively, given the QC
      Acceptance Criteria Table at the end of each of the determinative methods.
      If s and x for all  analytes of  interest meet the acceptance criteria, the
      system performance is  acceptable and analysis of actual samples can_begin.
      If any individual s exceeds  the precision  limit or any individual x falls
      outside  the  range   for accuracy,   then  the   system   performance  is
      unacceptable for that analyte.

            NOTE; The  large number  of analytes  in  each  of the  QC  Acceptance
                  Criteria Tables  present a  substantial probability that one or
                  more will  fail at  least one of the  acceptance  criteria when
                  all analytes of a given method are determined.

            8.6.6 When one or more of the analytes tested fail at least one of
      the  acceptance  criteria, the  analyst  should  proceed according  to  Step
      8.6.6.1 or 8.6.6.2.

                  8.6.6.1     Locate and correct the source of the problem and
            repeat  the  test   for  all   analytes  of  interest  beginning  with
            Step 8.6.2.

                  8.6.6.2     Beginning with Step 8.6.2,  repeat  the  test only
            for those analytes  that failed to meet criteria.  Repeated failure,
            however, will confirm a general problem with the measurement system.
            If this occurs, locate and  correct  the source of  the  problem and
            repeat  the  test  for  all  compounds  of  interest  beginning  with
            Step 8.6.2.

      8.7   The laboratory should,  on an  ongoing basis, analyze a reagent blank
and a matrix spiked duplicate for each analytical batch (up to a maximum of 20
samples/batch) to assess accuracy.  For soil and waste samples where detectable
amounts of organics are present, replicate samples may be appropriate in place
of spiked duplicates.   For laboratories analyzing  one to ten samples per month,
at least one spiked sample per month  is required.

            8.7.1 The. concentration  of  the spike  in  the sample  should  be
      determined as follows:

                  8.7.1.1     If,  as  in compliance monitoring, the concentration
            of  a  specific analyte in  the sample is  being checked  against  a
            regulatory concentration  limit,  the  spike should be at that limit,
            or 1 to 5 times  higher than the background concentration determined
            in Step 8.7.2, whichever  concentration would  be larger.

                  8.7.1.2     If the concentration  of a  specific  analyte  in  a
            water sample is not being checked against a limit  specific to that
            analyte,  the  spike  should  be at the  same concentration  as  the QC
            reference  sample  (Step  8.6.2}  or   1  to 5 times  higher than  the
            background  concentration  determined  in  Step  8.7.2,   whichever
            concentration would be larger.   For  other matrices, the recommended
            spiking concentration is  20 times the EQL.


                                  8000A - 11                        Revision 1
                                                                     July 1992

-------
            8.7.1.3     For semivolatile organics,  it may not be possible
      to determine the background  concentration  levels  prior to spiking
      (e.g. maximum holding  times will be exceeded).  If  this  is the case,
      the spike concentration should be (1) the regulatory concentration
      limit, if any; or,  if none (2) the larger of either 5 times higher
      than  the  expected background concentration or  the QC  reference
      sample  concentration   (Step   8.6.2).     For  other  matrices,  the
      recommended spiking concentration is 20 times the  EQL.

      8.7.2 Analyze  one  unspiked   and   one  spiked  sample  aliquot  to
determine percent recovery of each of the spiked compounds.

            8.7.2.1     Volatile  organics  -   Analyze   one  5-mL  sample
      aliquot  to determine  the background  concentration  (B)  of  each
      analyte.    If  necessary,  prepare  a  new  QC  reference  sample
      concentrate   (Step 8.6.1)    appropriate    for    the   background
      concentration in the  sample.   Spike a second  5-mL  sample aliquot
      with 10 nl of the QC reference sample concentrate  and analyze it to
      determine  the  concentration  after spiking  (A)  of  each  analyte.
      Calculate each percent recovery (p) as 100(A - B)%/T,  where T is the
      known true value of the spike.

            8.7.2.2     Semivolatile organics - Analyze one  sample aliquot
      (extract of  1-L  sample)  to determine the background concentration
      (B)  of  each analyte.    If necessary, prepare  a  new QC  reference
      sample  concentrate (Step  8.6.1)   appropriate  for the background
      concentration in the sample.  Spike a second 1-L sample aliquot with
      1.0 mL  of  the  QC  reference  sample concentrate and  analyze  it to
      determine  the  concentration  after spiking  (A)  of  each  analyte.
      Calculate each percent recovery (p) as 100(A - B)%/T,  where T  is the
      known true value of the spike.

      8.7.3 Compare the percent recovery (p) for each analyte in a water
sample  with the corresponding  criteria  presented in the  QC Acceptance
Criteria  Table  found  at the end of each of the determinative  methods.
These acceptance  criteria  were calculated  to  include  an  allowance for
error in  measurement of both  the  background  and  spike concentrations,
assuming a spike to background  ratio  of 5:1.  This error  will be accounted
for to the extent that the analyst's spike to background ratio approaches
5:1.    If  spiking  was performed at a  concentration  lower  than  the QC
reference sample concentration  (Step 8.6.2), the analyst  should use either
the  QC  acceptance  criteria presented   in  the  Tables,   or  optional  QC
acceptance criteria calculated for the specific  spike concentration.  To
calculate  optional  acceptance  criteria for the  recovery of  an  analyte:
(1)  Calculate accuracy  (x;)  using the  equation found  in  the  Method
Accuracy  and  Precision as a Function of Concentration  Table  (appears at
the   end   of  each   determinative  method),  substituting   the   spike
concentration  (T)  for  C;  (2) calculate overall_precision  (S')  using the
equation in the same  Table,  substituting  x' for x; (3) calculate the range
for recovery at the spike concentration  as (100x'/T)  ±  2.44(100S'/T)%.

      8.7.4 If any  individual  p falls  outside  the designated  range for
recovery,  that analyte  has failed  the acceptance  criteria.    A check
standard  containing  each  analyte   that  failed  the  criteria should be

                            800QA  -  12                        Revision  1
                                                               July  1992

-------
      analyzed as described in Step 8.8.

      8.8   If any analyte in a water sample fails the acceptance criteria for
recovery in Step 8.7, a QC reference standard containing each analyte that failed
should be prepared and analyzed.

      NOTE: The frequency for the required analysis of a QC reference standard
            will  depend upon the number of analytes being simultaneously tested,
            the complexity of the sample  matrix,  and the  performance  of the
            laboratory.  If the  entire list of analytes given in a method should
            be measured  in  the sample  in  Step 8.7,  the probability  that the
            analysis of a QC check standard will be required is high.  In this
            case, the QC  check  standard  should be routinely analyzed with the
            spiked sample.

            8.8.1 Preparation of the QC  check  sample  - For volatile organics,
      add 10 ML of  the  QC check sample  concentrate (Step 8.6.1  or 8.7.2) to 5
      ml of water.   For  semivolatile organics, add 1.0  ml of the QC check sample
      concentrate (Step  8.6.1 or  8.7.2)  to 1 L of water.  The  QC  check sample
      needs only to  contain  the analytes that failed  criteria in  the test in
      Step  8.7.    Prepare the  QC  check  sample  for  analysis  following  the
      guidelines given in Method 3500 (e.g. purge-and-trap, extraction, etc.).

            8.8.2 Analyze the QC  check  sample to  determine the  concentration
      measured (A)  of  each analyte.   Calculate  each  percent recovery  (ps) as
      100(A/T)%,  where T  is the true value of the standard  concentration.

            8.8.3 Compare the percent recovery (ps) for each analyte with the
      corresponding  QC  acceptance  criteria found in the appropriate Table in
      each of the methods. Only analytes that  failed the test in Step 8.7 need
      to be compared with these criteria.  If the recovery  of any such analyte
      falls outside  the  designated range,  the  laboratory performance for that
      analyte  is  judged  to   be  out of  control,  and  the  problem  should  be
      immediately identified and corrected.  The result for  that analyte  in the
      unspiked  sample  is suspect  and   may  not  be  reported  for  regulatory
      compliance purposes.

      8.9   As part  of'the QC program for the  laboratory,  method  accuracy for
each matrix studied should be  assessed and  records  should be maintained.  After
the analysis of five spiked samples  (of  the  same  matrix type)  as  in Step 8.7,
calculate the  average  percent recovery   (p) and the standard deviation of the
percent recovery (s  ).   Express the  accuracy assessment as a percent recovery
interval from  p  -  2s  to p +  2s .   If  p = 90%  and  s   = 10%,  for  example, the
accuracy interval is expressed as 70-110%.  Update the accuracy assessment for
each  analyte  on  a  regular basis  (e.g.  after  each five to ten  new accuracy
measurements).

      8.10  Calculate surrogate control   limits as follows:

            8.10.1       For each sample analyzed, calculate  the percent recovery
      of each  surrogate  in the sample.

            8.10.2      Calculate the average percent  recovery  (p) and standard
      deviation of  the percent recovery  (s)  for each of  the  surrogates when

                                  8000A  -  13                        Revision 1
                                                                     July 1992

-------
      surrogate data from 25 to 30 samples for each matrix is available.

            8.10.3      For  a  given  matrix,  calculate  the  upper and  lower
      control  limit for method performance  for  each  surrogate standard.   This
      should be done as follows:

            Upper Control Limit (UCL) » p + 3s
            Lower Control Limit (LCL) = p - 3s

            8.10.4      For  aqueous  and   soil   matrices,   these  laboratory
      established surrogate control  limits  should,  if  applicable,  be  compared
      with  the  control  limits in  Tables A and B of  Methods 8240 and  8270,
      respectively.   The limits  given  in these methods  are multi-laboratory
      performance based limits  for  soil  and  aqueous samples, and therefore, the
      single-laboratory limits established  in Step 8.10.3  should  fall  within
      those given in Tables A and B for these  matrices.

            8.10.5      If  recovery  is  not within limits,  the following  is
      required.

            •     Check  to  be  sure  there are  no  errors   in  calculations,
                  surrogate  solutions  and  internal  standards.   Also,  check
                  instrument performance.

            •     Recalculate the data and/or reanalyze the  extract if any of
                  the above checks reveal a problem.

            •     Reextract and reanalyze the  sample if none of the above are a
                  problem or flag the data as  "estimated concentration."

            8.10.6      At a minimum, each  laboratory  should update  surrogate
      recovery limits on a matrix-by-matrix basis, annually.

      8.11  It  is  recommended that  the  laboratory  adopt  additional  quality
assurance practices for use with  this method.   The specific practices  that are
most productive depend upon the needs of the  laboratory and the nature of the
samples.   Field duplicates  may  be  analyzed  to assess the precision of the
environmental  measurements.  When doubt exists  over the  identification of a peak
on the chromatogram, confirmatory techniques such as gas chromatography with a
dissimilar  column,  specific  element  detector, or mass spectrometer should be
used.   Whenever possible,  the laboratory  should analyze  standard  reference
materials and participate in relevant performance evaluation studies.


9.0   METHOD PERFORMANCE

      9.1   The  method  detection  limit  (MDL)   is  defined  as  the   minimum
concentration  of a  substance that  can  be  measured  and  reported  with  99%
confidence that the value is above zero.  The  MDL concentrations listed in the
referring analytical methods were  obtained  using  water.   Similar results were
achieved using representative wastewaters.  The MDL actually achieved in a given
analysis will  vary depending on instrument sensitivity and matrix effects.
                                  8000A - 14                        Revision 1
                                                                     July 1992

-------
      9.2   Refer to the determinative method  for specific method performance
information.
10.0  REFERENCES

1.     U.S.  EPA 40 CFR Part 136, "Guidelines Establishing Test Procedures for the
      Analysis of Pollutants Under the Clean Water Act; Final  Rule and Interim
      Final Rule and Proposed Rule," October 26, 1984.

2.     U.S.   EPA Contract  Laboratory Program,  Statement of  Work  for  Organic
      Analysis, July 1985, Revision.
                                  8000A  -  15                        Revision  1
                                                                     July  1992

-------
                                   METHOD  8000A
                               GAS CHROMATOGRAPHY
     Start
  7  1  Refer to
  determinative
  procedure for
   extraction
    procedure
 recommendation
  7  2  R.f.r  to
  determinative
  procedure  far
   cl.4nup and
   preparation
    proc.dur.
recommendations
                                   Internal Standard
                                                             External Standard
 7 4 1  C*tabli*h
 chromatographic
   condi 1 1 ons
  7431 Select
internal standards
  having behavior
    similar to
   compounds of
     in teres t
  7432 Prepar.
    c» 1 ibr it ion
    s tandards
  7433  Inj.ct
    calibration
    standards .
   calculate  RF
                        7434  Verify
                      •orking calibration
                       curve or  RF  each
                             day
  7421 Prepare
    calibration
standards for each
    compound of
     interest
  7422  Inject
    cal ibralion
standards,  prepare
 calibration curve
   or calculate
calibration factor
  7423  Verify
uorlting calibration
  curve each  day
                            7  S  Calculate
                           r.t.nIion  Iimi
                              •indo-i
                                      8000A -  16
                                                              Revision  1
                                                               July  1992

-------
                                       METHOD 8000A
                                         continued
                                              vol»ti!•»
     7 6.1 If
    neceaiary,
  screen aajaple*
  by M.thod 3810
     or 3820
  761 Introduce
 conpounda  into  CC
 by purgei-and-trap
or direct  injection
  (Method  S030)
                                        I
761  Introduce

 CC by direct
  injection
 7 6 *  Inject
 laraplea uting
 to 1venl fluih
  technique.
record  voluae
    7 6 5
Ooe> retponte
e»ceed linear
  range of
   • yatee,?
                               la peak
                              deteclion
                            prevented by
                            interference7
                                                      7 D S Dilute
                                                      extract and
                                                      reanalyie
   '  6  6 Do
    further
   c1eanup
                                                                                        7
                                                  68 Calibrate
                                                   *yatea
                                                  laaiediately
                                                  prior Io
                                                  analytea
                                                7 6 9 Eit.bU.h
                                                daily retention
                                                time iindo.i
                                                  for each
                                                  analyte
                         7 7 Perform
                       chrona tography
                           • ya tem
                       oaintenance . i f
                           needed
                                                7 8 Calculate
                                              concentration of
                                              each analyte  ujing
                                              appropriate formula
                                              for matnv and type
                                                of ilandard
                            Stop
                                        8000A  -  17
                                 Revision  1
                                  July  1992

-------
00
o
M*
o
03

-------
                                 METHOD 8010B

              HALOGENATED  VOLATILE ORGANICS  BY GAS CHROMATOGRAPHY
1.0   SCOPE AND APPLICATION

      1.1   Method 8010  is  used  to  determine  the  concentration  of  various
volatile  halogenated  organic  compounds.    The  following  compounds  can  be
determined by this method:
Appropriate Technique
Compound Name
Allyl chloride
Benzyl chloride
Bis(2-chloroethoxy)methane
Bis(2-chloroisopropyl ) ether
Bromoacetone
Bromobenzene
Bromodichloromethane
Bromoform
Bromomethane
Carbon tetrachloride
Chlorobenzene
Chloroethane
2-Chloroethanol
2-Chloroethyl vinyl ether
Chloroform
1-Chlorohexane
Chloromethane
Chloromethyl methyl ether
Chloroprene
4-Chlorotoluene
Di bromochl oromethane
1 , 2-Dibromo-3-chloropropane
Dibromomethane
1 , 2-Di chl orobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
l,4-Dichloro-2-butene
Di chl orodi fl uoromethane
1, 1-Dichloroethane
1,2-Dichloroethane
1, 1-Dichloroethene
trans-l,2-Dichloroethene
Di chl oromethane
1,2-Dichloropropane
l,3-Dichloro-2-propanol
cis-l,3-Dichloropropene
trans-l,3-Dichloropropene
Epichlorhydrin
CAS No.a
107-05-1
100-44-7
111-91-1
39638-32-9
598-31-2
108-86-1
75-27-4
75-25-2
74-83-9
56-23-5
108-90-7
75-00-3
107-07-03
110-75-8
67-66-3
544-10-5
74-87-3
107-30-2
126-99-8
106-43-4
124-48-1
96-12-8
74-95-3
95-50-1
541-73-1
106-46-7
764-41-0
75-71-8
75-34-3
107-06-2
75-35-4
156-60-5
75-09-2
78-87-5
96-23-1
10061-01-5
10061-02-6
106-89-8
Purge-and-Trap
b
PP
PP
b
PP
b
b
b
b
b
b
b
PP
b
b
pc
b
PP
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
PP
b
b
PP
Direct
Injection
b
b
pc
b
b
b
b
b
b
b
b
b
b
b
b
pc
b
pc
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
                                  8010B  -  1
    Revision 2
September 1994

-------
Compound Name
CAS No.'
  Appropriate Technique
                   Direct
Purge-and-Trap     Injection
Ethylene dibromide
Methyl iodide
1, 1,2,2-Tetrachloroethane
1,1,1 , 2-Tetrachl oroethane
Tetrachloroethene
1,1,1-Trichloroethane
1,1,2-Trichloroethane
Trichloroethene
Trichlorofluoromethane
1,2,3-Trichloropropane
Vinyl Chloride
a Chemical Abstract Servi
b Adequate response using
pp Poor purging efficiency
106-93-4
74-88-4
79-34-5
630-20-6
127-18-4
71-55-6
79-00-5
79-01-6
75-69-4
96-18-4
75-01-4
ces Registry Number
this technique
, resulting in high EQLs
b
PP
b
b
b
b
b
b
b
b
b



b
b
b
b
b
b
b
b
b
b
b



pc Poor chromatographic performance.
      1.2   Table 1 indicates compounds that may be  analyzed by this method and
lists the method detection limit for each compound in  organic-free reagent water.
Table 2 lists the estimated quantitation limit for other matrices.


2.0   SUMMARY OF METHOD

      2.1   Method  8010  provides  gas  chromatographic  conditions  for  the
detection of halogenated volatile organic compounds.  Samples can be introduced
into the GC  using direct injection or purge-and-trap (Method 5030).  Ground water
samples must be analyzed using Method 5030.   A temperature program is  used in the
gas chromatograph to separate the organic compounds.  Detection is achieved by
a electrolytic conductivity detector (HECD).

      2.2   The method provides an optional  gas  chromatographic column that may
be helpful  in resolving the analytes from co-eluting non-target  compounds and for
analyte confirmation.


3.0   INTERFERENCES

      3.1   Refer to Methods  5030 and  8000.

      3.2   Samples  can  be   contaminated  by diffusion  of volatile  organics
(particularly chlorofluorocarbons  and  methylene chloride) through  the  sample
container  septum  during  shipment  and  storage.   A  trip blank  prepared  from
organic-free reagent water and carried through sampling and subsequent storage
and handling can serve as a check on such contamination.
                                   8010B  -  2
                               Revision 2
                           September 1994

-------
4.0   APPARATUS AND MATERIALS

      4.1   Gas chromatograph

            4.1.1  Gas  chromatograph  -  analytical  system  complete  with  gas
      chromatograph suitable for on-column injections or purge-and-trap sample
      introduction and all required accessories,  including detector, analytical
      columns, recorder, gases, and syringes.   A  data  system for measuring peak
      heights and/or peak areas is recommended.

            4.1.2  Columns

                   4.1.2.1     Column 1 - 8 ft  x  0.1  in.  ID  stainless steel  or
            glass  column  packed with 1% SP-1000 on Carbopack-B  60/80 mesh  or
            equivalent.

                   4.1.2.2    Column 2 - 6 ft  x  0.1  in.  ID  stainless steel  or
            glass  column  packed with chemically bonded  n-octane  on Porasil-C
            100/120 mesh  (Durapak) or equivalent.

            4.1.3  Detector  -  Electrolytic  conductivity  (HECD).

      4.2   Sample  introduction  apparatus,   refer  to   Method  5030  for  the
appropriate equipment for sample introduction purposes.

      4.3   Syringes, 5 ml  Luerlok glass hypodermic and a 5 ml,  gas-tight with
shutoff valve.

      4.4   Volumetric  flask,   Class  A,  Appropriate  sizes  with  ground  glass
stoppers.

      4.5   Microsyringe, 10 and 25 /xL with a 0.006 in. ID needle (Hamilton 702N
or equivalent) and a 100 j^L.

      4.6   Analytical balance  - 0.0001  g.


5.0   REAGENTS

      5.1   Reagent  grade  chemicals  shall  be  used  in  all  tests.    Unless
otherwise  indicated,  it is  intended  that  all reagents  shall  conform to  the
specifications of the Committee on  Analytical  Reagents of the American Chemical
Society, where such  specifications are  available.   Other grades  may  be  used,
provided it is first ascertained that the reagent  is of sufficiently high purity
to permit its  use without lessening the accuracy  of the determination.

      5.2   Organic-free reagent water.   All  references to water in this method
refer to organic-free reagent water,  as defined in  Chapter One.

      5.3   Methanol, CH3OH.  Pesticide  quality or equivalent.  Store away from
other solvents.
                                  8010B  - 3                         Revision 2
                                                                September 1994

-------
      5.4   Stock standards - Stock solutions may be prepared from pure standard
materials or  purchased as  certified  solutions.   Prepare  stock  standards  in
methanol using assayed  liquids or gases,  as appropriate.  Because of the toxicity
of some  of  the organohalides,  primary dilutions of  these  materials should  be
prepared in a hood.

            5.4.1  Place  about  9.8  ml  of methanol in  a 10 ml tared ground glass
      stoppered volumetric  flask.   Allow the  flask to stand,  unstoppered,  for
      about 10 minutes until all  alcohol-wetted  surfaces have dried.  Weigh the
      flask to the nearest  0.0001 g.

            5.4.2  Add  the  assayed  reference material, as  described  below.

                   5.4.2.1     Liquids.  Using  a  100 /xL syringe, immediately add
            two or more drops of assayed reference material to the flask; then
            reweigh.   The  liquid must fall directly  into  the  alcohol  without
            contacting the  neck  of  the flask.

                   5.4.2.2     Gases.  To prepare  standards for any compounds
            that   boil   below   30°C   (e.g.   bromomethane,    chloroethane,
            chloromethane,   dichlorodifluoromethane,   trichlorofluoromethane,
            vinyl  chloride),  fill   a  5  ml valved  gas-tight syringe with  the
            reference  standard  to  the 5.0  ml mark.   Lower  the  needle  to 5  mm
            above  the methanol  meniscus.    Slowly  introduce  the  reference
            standard above  the  surface  of  the  liquid.  The heavy  gas  rapidly
            dissolves  in the methanol.  This may also be accomplished by using
            a  lecture  bottle equipped with  a Hamilton Lecture Bottle  Septum
            (#86600).  Attach  Teflon  tubing  to the  side-arm  relief valve  and
            direct a gentle stream  of gas into  the methanol meniscus.

            5.4.3  Reweigh,  dilute to volume,  stopper, and then mix by inverting
      the flask several times.   Calculate the concentration in  milligrams  per
      liter (mg/L)  from the  net gain in weight.  When  compound purity is  assayed
      to  be 96% or  greater,  the  weight may  be   used  without correction  to
      calculate the concentration of the stock standard.  Commercially prepared
      stock standards may be used at any concentration if they are certified  by
      the manufacturer or by an independent source.

            5.4.4  Transfer  the stock standard  solution  into a bottle  with  a
      Teflon lined screw-cap.  Store,  with  minimal  headspace,  at -10°C to -20°C
      and protect from light.

            5.4.5  Prepare  fresh  stock standards for  gases  weekly  or sooner  if
      comparison with check standards  indicates  a  problem.   Reactive compounds
      such as  2-chloroethyl  vinyl ether may need  to  be prepared more frequently.
      All other  standards  must  be  replaced  after six  months.   Both gas  and
      liquid standards must be monitored closely by  comparison  to  the  initial
      calibration curve  and by comparison  to QC  check  standards.    It  may  be
      necessary to  replace the standards more  frequently  if either check exceeds
      a 20% drift.

            5.4.6  Optionally calibration using  a certified gaseous mixture  can
      be accomplished  daily utilizing commercially  available gaseous  analyte


                                  8010B - 4                         Revision  2
                                                                September 1994

-------
      mixture  of bromomethane,  chloromethane,  chloroethane,  vinyl  chloride,
      dichlorodifluoromethane  and trichlorofluoromethane  in  nitrogen.  These
      mixtures  of documented  quality  are  stable  for as  long as  six  months
      without  refrigeration.  (VOA-CYL  III,  RESTEK  Corporation, Cat.  #20194 or
      equivalent).

      5.5   Secondary  dilution  standards.   Using   stock  standard  solutions,
prepare secondary  dilution standards  in  methanol,  as needed,  containing  the
compounds of interest,  either singly  or mixed together.  The secondary dilution
standards should be prepared at concentrations such  that the aqueous calibration
standards prepared in Sec.  5.6 will bracket  the working range of the analytical
system. Secondary dilution standards  should  be stored with minimal headspace for
volatiles  and  should  be  checked  frequently  for  signs  of  degradation  or
evaporation, especially just prior to preparing calibration standards from them.

      5.6   Calibration   standards.      Prepare   calibration   standards   in
organic-free reagent water from the secondary dilution of the stock standards,
at a minimum of five concentrations.   One of the concentrations should be at a
concentration  near,  but   above,  the  method detection limit.   The  remaining
concentrations  should  correspond  to  the  expected range of  the concentrations
found  in  real   samples or  should define  the  working range  of the GC.   Each
standard should contain each  analyte  for  detection by  this method (e.g. some or
all of the  compounds  listed  in Table 1 may be  included).  In  order to prepare
accurate aqueous standard  solutions, the following precautions must be observed.

            5.6.1  Do  not  inject more  than  20  /iL of  alcoholic standards into
      100 ml of water.

            5.6.2  Use  a   25   juL  Hamilton  702N  microsyringe  or  equivalent
      (variations  in   needle  geometry  will  adversely affect  the ability  to
      deliver  reproducible volumes of methanolic standards into water).

            5.6.3  Rapidly inject  the   alcoholic   standard  into  the  filled
      volumetric flask.  Remove the needle as fast as  possible after injection.

            5.6.4  Mix  aqueous standards by inverting the flask three times only.

            5.6.5  Fill the sample syringe from the standard solution contained
      in the expanded  area of the  flask  (do not  use any  solution  contained in
      the neck  of the  flask).

            5.6.6  Never  use  pipets to dilute  or transfer samples  or aqueous
      standards.

            5.6.7  Aqueous  standards are not  stable and should  be discarded after
      one hour,  unless properly  sealed and  stored.   The  aqueous standards  can
      be stored up to  24 hours, if held  in  sealed vials with zero headspace.

      5.7   Internal standards (if internal  standard  calibration is used) - To
use this approach, the analyst must select one or more internal standards that
are similar in  analytical   behavior to  the compounds  of  interest.   The analyst
must further demonstrate that  the measurement of the  internal  standard  is  not
affected by method or  matrix  interferences.  Because  of  these  limitations,  no


                                  8010B  - 5                        Revision 2
                                                                September 1994

-------
internal  standard  can be  suggested  that is  applicable  to all  samples.   The
compounds  recommended for use as  surrogate  spikes (Sec.  5.8)  have been used
successfully as internal  standards, because  of their generally  unique retention
times.

             5.7.1  Prepare  calibration   standards   at  a  minimum  of  five
      concentrations  for  each  analyte of interest as described in  Sec.  5.6.

             5.7.2  Prepare a spiking solution  containing  each  of the internal
      standards  using the procedures described  in  Sees.  5.4  and  5.5.   It  is
      recommended  that  the  secondary   dilution  standard  be  prepared at  a
      concentration  of  15 ng/juL  of each  internal  standard  compound.   The
      addition  of  10 p,i  of  this  standard to 5.0 ml of  sample or  calibration
      standard would  be equivalent to 30 M9A-

             5.7.3  Analyze each  calibration  standard  according to  Sec.  7.0,
      adding  10 /nL   of  internal   standard  spiking solution  directly  to  the
      syringe.

      5.8    Surrogate standards   -  The  analyst   should  monitor  both  the
performance  of  the analytical  system and the effectiveness of  the method  in
dealing  with  each  sample  matrix  by   spiking  each  sample,  standard,  and
organic-free reagent water blank with surrogate halocarbons.  A combination  of
bromochloromethane, bromochlorobenzene  and bromofluorobenzene is recommended  to
encompass  the  range  of temperature  program  used  in  this method.   From stock
standard solutions prepared as  in  Sec. 5.4, add a volume to give 750 /xg of each
surrogate to 45 ml  of organic-free reagent water contained in a  50 mL volumetric
flask, mix, and dilute to  volume  for  a concentration of 15 ng//iL.  Add 10 juL  of
this surrogate spiking solution directly  into the 5  ml syringe with every sample
and reference standard analyzed.  If the  internal  standard calibration procedure
is used, the surrogate compounds  may  be  added directly to the internal standard
spiking solution (Sec. 5.7.2).


6.0   SAMPLE COLLECTION,   PRESERVATION,  AND HANDLING

      6.1    See  the  introductory material to  this Chapter,  Organic Analytes,
Sec. 4.1.
7.0   PROCEDURE

      7.1   Volatile compounds are introduced into the gas chromatograph using
either direct  injection or purge-and-trap (Method 5030).  Method 5030 may be used
directly on ground  water samples or low-concentration  contaminated  soils and
sediments.   For medium-concentration  soils or sediments, methanolic extraction,
as described in Method  5030,  may  be necessary prior to purge-and-trap analysis.

      7.2   Gas  chromatographic  conditions  (Recommended)

            7.2.1 Column 1:

            Helium  flow  rate = 40 mL/min


                                  8010B  - 6                         Revision 2
                                                                September 1994

-------
            Temperature  program:
                   Initial  temperature  =   45°C, hold for 3 minutes
                   Program  =               45°C to 220°C at 8°C/min
                   Final  temperature =     220°C, hold for 15 minutes.

            7.2.2  Column 2:

            Helium flow  rate = 40 mL/min
            Temperature  program:
                   Initial  temperature  =   50°C, hold for 3 minutes
                   Program  =               50°C to 170°C at 6°C/min
                   Final  temperature =     170°C, hold for 4 minutes.

      7.3   Calibration.  The procedure for internal or external calibration may
be used.  Refer to Method 8000 for a  description of each of these procedures. Use
Table 1  and Table 2 for guidance on selecting the lowest point on the calibration
curve.

            7.3.1  Calibration must take place using the same sample introduction
      method that will be used to analyze actual samples  (see Sec. 7.4.1).

      7.4   Gas chromatographic analysis

            7.4.1  Introduce  volatile compounds into the gas  chromatograph using
      either Method 5030 (purge-and-trap) or  the  direct injection method (see
      Sec. 7.4.1.1).   If the internal  standard calibration technique is used,
      add 10 n\. of internal  standard to the sample  prior  to purging.

                   7.4.1.1     In very limited applications (e.g. aqueous process
            wastes) direct  injection  of  the sample onto  the  GC  column  with a
            10 juL  syringe may be appropriate.   The  detection limit is very high
            (approximately 10,000 M9/L) therefore,  it is only permitted where
            concentrations in excess of 10,000 M9/L are expected or for water-
            soluble compounds that do not  purge.  The  system must be calibrated
            by direct  injection (bypassing the purge-and-trap device).

            7.4.2  Method 8000 provides instructions on the analysis sequence,
      appropriate  dilutions,  establishing daily  retention time  windows,  and
      identification criteria.   Include a  mid-concentration  standard after each
      group of 10 samples in the analysis sequence.

            7.4.3  Table  1  summarizes  the  estimated retention times  on the two
      columns for a number of organic  compounds  analyzable using this method.
      An example of the  separation  achieved by Column  1 is  shown in Figure 1.

            7.4.4  Record the sample volume purged  or injected and the resulting
      peak sizes (in area units or  peak heights).

            7.4.5  Refer  to  Method  8000  for  guidance  on  calculation  of
      concentration.

            7.4.6  If analytical interferences are suspected, or for the purpose
      of confirmation, analysis using the second GC column  is recommended.


                                  8010B - 7                         Revision 2
                                                                September 1994

-------
             7.4.7  If the response for  a  peak is off-scale,  i.e.,  beyond the
      calibration range of the standards,  prepare a dilution of the sample with
      organic-free reagent water.  The  dilution  must  be performed on a second
      aliquot of the sample which has been properly sealed and stored prior to
      use.
8.0   QUALITY CONTROL

      8.1    Refer  to  Chapter One for specific quality  control  procedures and
Method 8000 for gas chromatographic procedures.  Quality control to ensure the
proper operation of the purge-and-trap device is covered in Method 5030.

      8.2    Quality  control  required to  validate  the GC  system  operation is
found in Method 8000.

             8.2.1  The  quality control check  sample  concentrate (Method 8000)
      should contain each analyte of  interest at a concentration of 10 mg/L in
      methanol.

             8.2.2  Table 3 indicates the calibration and QC acceptance criteria,
      for water  samples,  for this method.  Table  4  gives  method  accuracy and
      precision  as functions  of concentration,  for water  samples,   for  the
      analytes  of  interest.    The  contents of both  Tables should be  used to
      evaluate a laboratory's  ability to  perform  and generate acceptable data
      by this method.

      8.3    Calculate  surrogate  standard  recovery  on all  samples,  blanks,  and
spikes.   Determine  if recovery is  within limits (limits established by performing
QC procedure outlined in Method 8000).

             8.3.1  If  recovery  is not  within limits, the following is required:

                   •     Check  to  be  sure  that  there   are  no   errors  in
             calculations, surrogate  solutions and internal standards.   Also,
             check  instrument performance.

                   •      Recalculate  the data and/or  re-analyze the  sample if
             any  of the above checks reveal a problem.

                   •     Re-extract  and  re-analyze  the sample  if  none  of  the
             above  are a problem  or flag the data as "estimated concentration".


9.0   METHOD PERFORMANCE

      9.1    This method was tested by 20 laboratories  using organic-free reagent
water, drinking  water,  surface water,  and three industrial wastewaters spiked at
six concentrations over  the  range 8.0-500 /^g/L.   Single  operator  precision,
overall  precision,  and method accuracy were found to  be directly related to the
concentration of the  analyte,  and essentially  independent of the sample matrix.
Linear equations to describe  these relationships are presented in  Table 4.
                                  8010B  - 8                         Revision 2
                                                                September 1994

-------
      9.2   The accuracy and precision  obtained will be determined by the sample
matrix, sample introduction technique,  and by the calibration procedure used.

      9.3   The method detection limits reported  in Table 1 were generated under
optimum analytical conditions by an Agency contractor  (Ref. 6) as guidance, and
may not be readily achievable by all laboratories at all times.


10.0  REFERENCES

1.    Bellar, T.A.; Lichtenberg, J.J. jh  Amer. Water Works Assoc. 1974, 66(12),
      pp. 739-744.

2.    Bellar,  T.A.;  Lichtenberg,  J.J.,  Semi-Automated Headspace  Analysis of
      Drinking  Waters  and  Industrial  Waters for Purgeable  Volatile Organic
      Compounds, Measurement of Organic Pollutants in  Water and Wastewater; Van
      Hall, Ed.; ASTM STP 686, pp  108-129, 1979.

3.    "Development and Application of Test Procedures for Specific Organic Toxic
      Substances  in  Wastewaters: Category 11  -  Purgeables and  Category  12  -
      Acrolein,  Acrylonitrile,   and Dichlorodifluoromethane";  report  for  EPA
      Contract 68-03-2635.

4.    U.S. EPA 40 CFR Part 136, "Guidelines Establishing Test Procedures for the
      Analysis of Pollutants Under the Clean Water Act: Final  Rule and Interim
      Final Rule  and Proposed Rule", October 26,  1984.

5.    "EPA  Method Validation Study 23,  Method  601  (Purgeable  Halocarbons)";
      report for  EPA Contract 68-03-2856.

6.    Gebhart, J.E., S.V. Lucas, S.J.  Naber,  A.M. Berry, T.H.  Danison and H.M.
      Burkholder,  "Validation of SW-846  Methods  8010,  8015,  and 8020"; Report
      for  EPA  Contract  68-03-1760,  Work Assignment  2-15;    US   EPA,  EMSL-
      Cincinnati, 1987.
                                   8010B  -  9                         Revision 2
                                                                September 1994

-------
                       TABLE  1.
CHROMATOGRAPHIC CONDITIONS AND METHOD DETECTION LIMITS
           FOR HALOGENATED VOLATILE ORGANICS
Compound
Ally! chloride*^
Benzyl chloride*'0
Bis(2-chloroethoxy)methane*
Bis(2-chloroisopropyl ) ether*
Bromobenzene
Bromodichloromethane
Bromoform*
Bromomethane*
Carbon tetrachloride*
Chi oroacet aldehyde*
Chlorobenzene*
Chl oroethane
Chloroform*
1-Chlorohexane
2-Chloroethyl^vinyl ether*
Chloromethane*
Chloromethyl methyl ether*
4-Chlorotoluene
Dibromochloromethane
1 , 2-Dibromo-3-chl oropropane*
Dibromomethane*
1 , 2-Di chl orobenzene*
1, 3 -Di chlorobenzene*
1 ,4-Dichl orobenzene*
l,4-Dichloro-2-butene*
Di chl orodi f 1 uoromethane*'d
1,1-Dichloroethane*
1,2-Dichloroethane*
1,1-Dichloroethene*
trans - 1 , 2-Di chl oroethene*
Dichloromethane*
1, 2-Di chl oropropane*
trans - 1 , 3-Di chl oropropene*
Ethylene dibromide
1,1,2 , 2-Tetrachl oroethane*
1,1, 1, 2-Tetrachl oroethane*
Tetrachl oroethene*
1 , 1 , 1 -Tri chl oroethane^
1, 1, 2 -Tri chl oroethane*
CAS
Registry
Number
107-05-1
100-44-7
111-91-1
39638-32-9
108-86-1
75-27-4
75-25-2
74-83-9
56-23-5
107-20-0
108-90-7
75-00-3
67-66-3
544-10-5
110-75-8
74-87-3
107-30-2
106-43-4
124-48-1
96-12-8
74-95-3
95-50-1
541-73-1
106-46-7
764-41-0
75-71-8
75-34-3
107-06-2
75-35-4
156-60-5
75-09-2
78-87-5
10061-02-5
106-93-4
79-34-5
630-20-6
127-18-4
71-55-6
79-00-5
Retention Time
(minutes)
Column 1 Column 2
10.17
30.29
38.60
34.79
29.05
15.44
21.12
2.90
14.58
(b)
25.49
5.18
12.62
26.26
19.23
1.40
8.88
34.46
18.22
28.09
13.83
37.96
36.88
38.64
23.45
3.68
11.21
13.14
10.04
11.97
7.56
16.69
16.97e
19.59
23.12
21.10
23.05
14.48
18.27
(b)
(b)
(b)
(b)
(b)
14.62
19.17
7.05
11.07
(b)
18.83
8.68
12.08
(b)
(b)
5.28
(b)
(b)
16.62
(b)
14.92
23.52
22.43
22.33
(b)
(b)
12.57
15.35
7.72
9.38
10.12
16.62
16.60
(b)
(b)
21.70
14.97
13.10
18.07
Method
Detection
Limit3
(M9/L)
(b)
(b)
(b)
(b)
(b)
0.002
0.020
0.030
0.003
(b)
0.001
0.008
0.002
(b)
0.130
0.010
(b)
(b)
(b)
0.030
(b)
(b)
(b)
(b)
(b)
(b)
0.002
0.002
0.003
0.002
(b)
(b)
0.340
(b)
0.010
(b)
0.001
0.003
0.007
                      8010B - 10
    Revision 2
September 1994

-------
                                   TABLE 1.
                                   Continued
   Compound
                             CAS
                             Registry
                             Number
  Retention Time
     (minutes)
Column 1   Column
          Method
          Detection
          Limit9
          (M9/L)
Trichloroethene*
Trichlorofluoromethane*
1,2,3-Trichloropropane*
Vinyl Chloride*
                              79-01-6
                              75-69-4
                              96-18-4
                              75-01-4
 17.40
  9.26
 22.95
  3.25
13.12
 (b)
 (b)
 5.28
0.001
 (b)
 (b)
0.006
a =
b =
* =
c =
d =

e =
Using purge-and-trap method (Method 5030). See Sec. 9.3.
Not determined
Appendix VIII compounds
Demonstrated very erratic results when tested by purge-and-trap
See  Sec.  4.10.2  of Method  5030 for  guidance  on  selection  of  trapping
material
Estimated retention time
                                   TABLE 2.
             DETERMINATION OF ESTIMATED QUANTITATION LIMITS (EQL)
                             FOR VARIOUS MATRICES3
               Matrix
                                               Factor
               Ground water                             10
               Low-concentration soil                   10
               Water miscible liquid waste             500
               High-concentration soil and sludge     1250
               Non-water miscible waste               1250
               EQL = [Method detection limit (see Table 1)] X [Factor found in
               this table].  For  non-aqueous samples,  the  factor is on  a wet-
               weight basis. Sample EQLs are highly matrix-dependent.  The EQLs
               listed herein are  provided  for  guidance and may  not always be
               achievable.
                                  8010B - 11
                                                                Revision 2
                                                            September 1994

-------
                             TABLE 3.
             CALIBRATION AND QC ACCEPTANCE  CRITERIA3
Range Limit
for Q for S
Analyte
Bromodi chloromethane
Bromoform
Bromomethane
Carbon tetrachloride
Chlorobenzene
Chloroethane
2-Chloroethylvinyl ether
Chloroform
Chloromethane
Di bromochl oromethane
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Di chlorobenzene
1,1-Dichloroethane
1,2-Dichloroethane
1,1-Dichloroethene
trans-l,2-Dichloroethene
Dichloromethane
1,2-Dichloropropane
cis-l,3-Dichloropropene
trans-l,3-Dichloropropene
1,1,2 , 2-Tetrachl oroethane
Tetrachloroethene
1,1,1-Trichloroethane
1,1,2-Trichloroethane
Trichloroethene
Tri chl orof 1 uoromethane
Vinyl chloride
(M9/L) (M9/L)
15.2-24.8
14.7-25.3
11.7-28.3
13.7-26.3
14.4-25.6
15.4-24.6
12.0-28.0
15.0-25.0
11.9-28.1
13.1-26.9
14.0-26.0
9.9-30.1
13.9-26.1
16.8-23.2
14.3-25.7
12.6-27.4
12.8-27.2
15.5-24.5
14.8-25.2
12.8-27.2
12.8-27.2
9.8-30.2
14.0-26.0
14.2-25.8
15.7-24.3
15.4-24.6
13.3-26.7
13.7-26.3
4.3
4.7
7.6
5.6
5.0
4.4
8.3
4.5
7.4
6.3
5.5
9.1
5.5
3.2
5.2
6.6
6.4
4.0
5.2
7.3
7.3
9.2
5.4
4.9
3.9
4.2
6.0
5.7
Q = Concentration measured in QC check sampl
Range
for x
(M9/L)
10.7-32.0
5.0-29.3
3.4-24.5
11.8-25.3
10.2-27.4
11.3-25.2
4.5-35.5
12.4-24.0
D-34.9
7.9-35.1
1.7-38.9
6.2-32.6
11.5-25.5
11.2-24.6
13.0-26.5
10.2-27.3
11.4-27.1
7.0-27.6
10.1-29.9
6.2-33.8
6.2-33.8
6.6-31.8
8.1-29.6
10.8-24.8
9.6-25.4
9.2-26.6
7.4-28.1
8.2-29.9
e, in M9/L.
S = Standard deviation of four recovery measurements, in
x = Average recovery
P, Ps = Percent recovery
D = Detected; result
a r  •; •§• o *« i r» -Pv»/->m At\ rco Dn
Range
P>oPs

42-172
13-159
D-144
43-143
38-150
46-137
14-186
49-133
D-193
24-191
D-208
7-187
42-143
47-132
51-147
28-167
38-155
25-162
44-156
22-178
22-178
8-184
26-162
41-138
39-136
35-146
21-156
28-163

M9/L.
for four recovery measurements, in /ng/L.
measured.
must be greater
n+ 1 TC fnv Mat hn.



than zero.
A Cm on
f\ lilQV^Q ^ol^Mll
3 "H or! srciiminn
a QC check sample concentration of 20 p.g/1.
                            8010B - 12
    Revision 2
September 1994

-------
                                TABLE 4.
      METHOD ACCURACY AND PRECISION AS FUNCTIONS OF CONCENTRATION3
Analyte
Bromodichl oromethane
Bromoform
Bromomethane
Carbon tetrachloride
Chlorobenzene
Chloroethane
2-Chloroethyl vinyl etherb
Chloroform
Chloromethane
Di bromochl oromethane
1 ,2-Dichlorobenzene
1 ,3-Dichlorobenzene
1 ,4-Dichlorobenzene
1 ,1-Dichloroethane
1 ,2-Dichloroethane
1 ,1-Dichloroethene
trans-l,2-Dichloroethene
Dichl oromethane
l,2-Dichloropropaneb
cis-l,3-Dichloropropeneb
trans-1 ,3-Dichloropropeneb
1 , 1,2, 2-Tetrachloroethane
Tetrachloroethene
1,1,1-Tri chloroethane
1, 1, 2 -Tri chloroethane
Trichloroethene
Trichlorofl uoromethane
Vinyl chloride
Accuracy, as
recovery, x'
(M9A)
1.12C-1.02
0.96C-2.05
0.76C-1.27
0.98C-1.04
l.OOC-1.23
0.99C-1.53
l.OOC
0.93C-0.39
0.77C+0.18
0.94C+2.72
0.93C+1.70
0.95C+0.43
0.93C-0.09
0.95C-1.08
1.04C-1.06
0.98C-0.87
0.97C-0.16
0.91C-0.93
l.OOC
l.OOC
l.OOC
0.95C+0.19
0.94C+0.06
0.90C-0.16
0.86C+0.30
0.87C+0.48
0.89C-0.07
0.97C-0.36
Single analyst
precision, s '
(M9/L)
0.11X+0.04
0.12X+0.58
0.28X+0.27
0.15X+0.38
0.15X-0.02
0.14X-0.13
0.20X
0.13X+0.15
0.28X-0.31
0.11X+1.10
0.20X+0.97
0.14X+2.33
0.15X+0.29
0.08X+0.17
0.11X+0.70
0.21X-0.23
0.11X+1.46
0.11X+0.33
0.13X
0.18X
0.18X
0.14X+2.41
0.14X+0.38
0.15X+0.04
0.13X-0.14
0.13X-0.03
0.15X+0.67
0.13X+0.65
Overall
precision,
S' (Mg/L)
0.20X+1.00
0.21X+2.41
0.36X+0.94
0.20X+0.39
0.18X+1.21
0.17X+0.63
0.35X
0.19X-0.02
0.52X+1.31
0.24X+1.68
0.13X+6.13
0.26X+2.34
0.20X+0.41
0.14X+0.94
0.15X+0.94
0.29X-0.04
0.17X+1.46
0.21X+1.43
0.23X
0.32X
0.32X
0.23X+2.79
0.18X+2.21
0.20X+0.37
0.19X+0.67
0.23X+0.30
0.26X+0.91
0.27X+0.40
x' =  Expected recovery for one or more measurements of a sample containing
      a concentration of C, in M9/L-
sr'=  Expected  single  analyst  standard deviation  of  measurements  at  an
      average concentration of x,  in ng/L.
S' =  Expected  interlaboratory  standard deviation  of  measurements  at  an
      average concentration found  of x, in  M9/L-
C  =  True value for the concentration, in  M9/L-
X  =  Average  recovery  found  for  measurements  of  samples  containing  a
      concentration of C, in M9/L.

a From 40 CFR Part 136 for Method  601.
b Estimates  based upon  the performance  in a single laboratory.
                               8010B - 13
    Revision 2
September 1994

-------
                       FIGURE  1.
GAS  CHROMATOGRAM OF HALOGENATED VOLATILE ORGANICS
                                 CoI urn:   1% SP-1000 on Carbopack-B
                                 Program:  45"C-3 Minutes, 8'C/Minute to 220CC
                                 Detector: Hall 700-A Electrolytic Conductivity
  •    10   12   M   l«   It    20    n


                     METCWTION TMC MIIIUTHI
                       8010B - 14
     Revision 2
September  1994

-------
                                      METHOD  8010B
             HALOGENATED VOLATILE  ORGANICS BY GAS CHROMATOGRAPHY
        Start
7.1  Introduce compounds
into gas chromatograph
  by direct injection or
    purge-end-trap
     (Method 5030)
      7.2 Set gas
    chromatograph
      condition.
     7.3 Calibrate
 (refer to Method 8000)
    7.4.1  Introduce
  volatile  compounds
 into gas chromatograph
  by purge-and-trap or
    direct injection.
  7.4.2 Follow Method
   8000  for analysis
     sequence, etc.
7.4.4 Record volume
 purged or injected
  and peak sizes.
    7.4.5 Calculate
    concentration
(refer to Method 8000)
      7.4.6 Are
      analytical
     interferences
     suspected?
                                             7.4.7 Is
                                           response for
                                             a peak
                                            off-scale?
                                                                    7.4.6 Analyze using
                                                                    second GC column.
                               7.4.7 Dilute second
                                aliquot of sample.
                                       8010B -  15
                                              Revision 2
                                         September  1994

-------

-------
                                  METHOD 8011

              1.2-DIBROMOETHANE AND 1,2-DIBROMO-3-CHLOROPROPANE
                   BY  MICROEXTRACTION AND GAS  CHROMATOGRAPHY
1.0   SCOPE AND APPLICATION

      1.1   This method  is  applicable  to  the determination of  the  following
compounds in drinking water and ground water:
      Compound Name                                   CAS No."


      1,2-Dibromoethane (EDB)                         106-93-4
      l,2-Dibromo-3-chloropropane (DBCP)               96-12-8


      8   Chemical  Abstract Services  Registry Number.

      1.2   For compounds and matrices other than those listed in Section 1.1,
the laboratory  must demonstrate  the usefulness of  the method  by  collecting
precision  and  accuracy  data  on  actual  samples  and  provide  qualitative
confirmation of results by gas chromatography/mass spectrometry (GC/MS).

      1.3   The experimentally determined method detection limits (MDL) for EOB
and DBCP were  calculated to be  0.01  M9/L.   The method has been shown  to  be
useful for  these analytes over a concentration  range of approximately  0.03 to 200
/ig/L.  Actual  detection limits are highly dependent upon the characteristics of
the gas chromatographic system, sample matrix, and calibration.

      1.4   This method  is  restricted to  use by or under  the Supervision  of
analysts experienced in the use of gas chromatography and in the interpretation
of gas  chromatograms.  Each analyst  must demonstrate the ability to  generate
acceptable results with this method using the procedure described in Section 8.2.

      1.5   1,2-Dibromoethane   and    l,2-Dibromo-3-chloropropane    have   been
tentatively classified  as known or  suspected  human  or  mammalian carcinogens.
Pure standard materials and stock standard solutions of these compounds should
be handled  in  a hood.  A NIOSH/MESA approved toxic gas respirator should be worn
when the analyst handles high concentrations of these toxic compounds.


2.0   SUMMARY OF METHOD

      2.1   Thirty five ml of sample are extracted with 2 ml of hexane.  Two  jiL
of the  extract are  then injected  into a gas  chromatograph equipped  with  a
linearized electron capture detector for separation and analysis. Aqueous matrix
spikes are  extracted and analyzed in an identical manner as the  samples in order
to compensate for possible extraction losses.

      2.2   The extraction  and analysis  time  is  30  to 50  minutes  per  sample

                                   8011  -  1                         Revision 0
                                                                     July 1992

-------
depending upon the analytical conditions chosen.  See Table 1 and Figure 1.

      2.3   Confirmatory evidence is obtained using a different column (Table 1).


3.0   INTERFERENCES

      3.1   Impurities  contained  in the  extracting solvent  (hexane)  usually
account for the majority of the analytical problems.  Reagent blanks should be
analyzed for each new bottle  of hexane before use.  Indirect dally checks on the
hexane are obtained by monitoring the reagent blanks.  Whenever an Interference
is noted in the method or instrument blank, the  laboratory  should reanalyze the
hexane.   Low  level  interferences  generally can  be  removed by  distillation or
column chromatography, however,  it  is generally  more economical to obtain a new
source of hexane solvent.  Interference-free hexane is defined as containing less
than 0.01 pg/L of the analytes.  Protect interference-free  hexane by storing it
in an area known to be free of organochlorine solvents.

      3.2   Several  Instances of  accidental  sample  contamination have  been
attributed to  diffusion  of volatile organics through the septum seal  Into the
sample bottle during shipment and storage.  Trip blanks must be used to monitor
for this problem.

      3.3   This liquid/liquid extraction technique extracts a wide boiling range
of non-polar  organic compounds  and,  in addition,  extracts some  polar organic
compounds.

      3.4   EDB at low concentrations may be masked  by very high concentrations
of dibromochloromethane (DBCM), a common chlorinated drinking water contaminant,
when using the confirmation column.


4.0   APPARATUS AND MATERIALS

      4.1   Microsyringe  - 10, 25, and  100 pi with a 2 in. x 0.006 1n. needle
(Hamilton 702N or equivalent).

      4.2   Gas Chromatograph

            4.2.1 The GC  must be capable of temperature programming and should
      be  equipped with  a linearized electron capture detector  and a capillary
      column  splitless  injector.

            4.2.2 Columns

                  4.2.2.1     Column  A  -  0.32  mm  ID x  30  m  fused  silica
            capillary with dimethyl silicone mixed  phase  (Durawax-DX 3, 0.25 urn
            film, or equivalent).

                  4.2.2.2     Column B (confirmation column) - 0.32 mm  ID x 30 m
            fused silica capillary  with methyl polysiloxane phase (DB-1, 0.25 ^m
            film, or equivalent).

      4.3   Volumetric  flasks, Class A  -  10 mL.

                                   8011 - 2                         Revision 0
                                                                     July 1992

-------
      4.4   Glass bottles - 15 ml, with Teflon lined screw caps or crimp tops.

      4.5   Analytical  balance - 0.0001 g.

      4.6   Graduated cylinder - 50 ml.

      4.7   Transfer pipet.


5.0   REAGENTS

      5.1   Reagent grade chemicals shall be used in  all tests. Unless otherwise
indicated, it is intended that all reagents shall conform to the specifications
of the Committee on Analytical  Reagents of the American Chemical Society, where
such specifications are available. Other grades may be used,  provided it is first
ascertained that the reagent  is of sufficiently  high  purity  to permit  its use
without lessening the accuracy of the determination.

      5.2   Organic-free reagent water - All references to water in this method
refer to organic-free reagent water,  as defined in Chapter One.

      5,3   Hexane, C6HU - UV grade (Burdick and Jackson 1216 or equivalent).

      5.4   Methyl  alcohol, CH3OH  - Demonstrated  to  be free of analytes.

      5.5   Sodium chloride,  NaCl  - Pulverize a batch  of NaCl and place 1t in a
muffle  furnace  at  room  temperature.   Increase  the temperature to  400°C for
30 minutes.  Store in a capped bottle.

      5.6   1,2-Dibromoethane  (99%),  C2H4Br2,  (Aldrich Chemical   Company,  or
equivalent).

      5.7   l,2-Dibromo-3-chloropropane  (99.4%),  C3H5Br2Cl,   (AMVAC  Chemical
Corporation,  Los Angeles, California,  or  equivalent).

      5.8   Stock  standards  - These   solutions may  be purchased as  certified
solutions or prepared from pure standards  using the following procedures:

            5.8.1 Place  about  9.8 ml of  methanol  into a  10 ml ground  glass
      stoppered volumetric flask.  Allow  the flask  to stand,  unstoppered, for
      about 10 minutes  and weigh to the nearest 0.0001 g.

            5.8.2 Use a  25 pi syringe and  immediately  add two or more  drops
      (= 10 pi)  of standard  to  the  flask.   Be  sure  that the  standard  falls
      directly into the alcohol without contacting the neck of the  flask.

            5.8.3 Reweigh, dilute to volume, stopper,  and then mix by inverting
      the flask several  times.  Calculate the  concentration  in milligrams per
      liter (mg/L)  from  the net gain in weight.  When compound purity is assayed
      to  be  96% or  greater,  the  weight  may  be used  without correction  to
      calculate the concentration of  the  stock standard.

            5.8.4 Store stock standards in 15 ml bottles equipped  with Teflon
      lined screw-caps  or crimp tops.   Stock standards are stable  for at least

                                   8011 -  3                          Revision 0
                                                                     July 1992

-------
      four weeks when stored at 4°C and away from light.

      5.9   Intermediate  standard   -   Use   stock  standards  to  prepare  an
intermediate standard that contains  both analytes in methanol. The intermediate
standard should  be  prepared at a concentration that  can  be  easily diluted to
prepare aqueous calibration  standards that will bracket the working concentration
range.   Store  the  intermediate  standard  with minimal  headspace  and  check
frequently  for  signs of deterioration  or evaporation,  especially just before
preparing calibration standards.   The storage time described for  stock standards
also applies to the  intermediate  standard.

      5.10  Quality control  (QC) reference sample -  Prepare a QC reference sample
concentrate at 0.25 mg/L of  both analytes from standards from  a different source
than the standards used  for the stock  standard.

      5.11  Check  standard  -  Add an  appropriate  volume of the intermediate
standard to an aliquot of organic-free reagent water in  a  volumetric flask.  Do
not add more than  20 /iL of  an  alcoholic intermediate standard to the water or
poor precision will  result.  Use a  25 juL microsyringe  and rapidly inject the
alcoholic  intermediate  standard   into  the expanded area  of  the almost filled
volumetric flask.  Remove the needle as quickly as possible after injection.  Mix
by inverting the  flask several times.  Discard the contents contained in  the neck
of the flask.  Aqueous  calibration  standards  should be  prepared every 8 hours.


6.0   SAMPLE COLLECTION, PRESERVATION, AND STORAGE

      6.1   See  the  introductory  material  to  this  chapter,  Organic Analytes,
Section 4.1.
7.0   PROCEDURE

      7.1   Recommended Chromatographic Conditions

      Two  gas  chromatography columns are  recommended.   Column A  is a highly
efficient column that provides separations for EDB and DBCP without interferences
from trihalomethanes.  'Column A should be used as the  primary analytical column
unless routinely  occurring  analytes  are  not adequately resolved.   Column B is
recommended for  use  as a  confirmatory  column when GC/MS  confirmation is not
available.  Retention times  for EDB  and DBCP  on these columns are presented in
Table 1.

      Column A:
       Injector  temperature:                     200 C.
       Detector  temperature:                     290°C.
       Carrier gas  (Helium)  Linear velocity:     25 cm/sec.
       Temperature  program:
             Initial  temperature:                40°C, hold for 4 min.
             Program:                            40°C to 190°C at 8°C/min.
             Final  temperature:                  190°C,  hold   for  25  min.,  or
                                                until   all   expected  analytes
                                                have  eluted.

                                   8011  -  4                          Revision 0
                                                                      July 1992

-------
See Figure 1 for a sample chromatogram and Table  1 for retention data.

Column B:

Injector temperature:                     200°C.
Detector temperature:                     290°C.
Carrier gas (Helium) Linear velocity:     25 cm/sec.
Temperature program:
      Initial temperature:                40°C, hold for 4 min.
      Program:                            40°C to 270°C  at 10°C/min.
      Final temperature:                  270°C,  hold  for  10 min.,  or
                                          until   all  expected  analytes
                                          have eluted.

See Table 1 for retention data.

7.2   Calibration

      7.2.1 Prepare  at  least  five calibration  standards.    One  should
contain EDB and DBCP at a concentration near,  but greater than, the method
detection limit  (Table  1) for each  compound.   The others  should  be at
concentrations  that bracket  the  range  expected in  the samples.    For
example,  if the HDL is  0.01  M9/L.  and a  sample  expected  to  contain
approximately 0.10 /ug/L is to be analyzed, aqueous calibration standards
should be prepared at concentrations of 0.03  /ug/L, 0.05 M9/U 0.10  M9/U
0.15 M9/U and 0.20  /ig/L.

      7.2.2 Analyze each calibration standard and tabulate peak height or
area  response versus  the concentration in  the standard.   Prepare  a
calibration  curve  for each  compound.   Alternatively,  if  the  ratio of
response  to  concentration (calibration  factor)  is  a constant  over the
working  range (<  10% relative  standard deviation),  linearity can be
assumed and the average ratio or calibration factor can be used in  place
of a calibration curve.

7.3   Sample preparation

      7.3.1 Remove  samples and standards from storage  and allow them to
reach room temperature.

      7.3.2 For  samples  and  field blanks contained  in  40  ml  bottles,
remove the  container cap.   Discard a 5  ml volume using  a 5 ml transfer
pipet.  Replace the container cap and weigh the container with  contents to
the  nearest  0.1  g  and  record  this weight for subsequent sample  volume
determination.

      7.3.3 For  calibration  standards,  check standards,  QC reference
samples,  and blanks, measure  a 35  mi  volume  using  a  50  ml graduated
cylinder  and transfer it to a 40 ml sample container.

7.4   Extraction

      7.4.1 Remove  the container cap  and add 7 g  of NaCl  to all  samples.
                             8011  - 5                         Revision 0
                                                               July 1992

-------
      7.4.2 Recap the sample container and dissolve the NaCl  by  shaking by
hand for about 20 seconds.

      7.4.3 Remove the  cap and using  a transfer  pi pet,  add  2.0  ml of
hexane.  Recap and shake vigorously by hand for 1  minute.  Allow the water
and  hexane phases  to separate.    If  stored  at  this  stage,  keep  the
container upside down.

      7.4.4 Remove the  cap and carefully  transfer a  sufficient amount
(0.5-1.0 ml)  of the hexane layer  into  a vial using  a disposable glass
pipet.

      7.4.5 Transfer  the  remaining hexane  phase,  being careful  not to
include any of the water phase, into a second vial.  Reserve this second
vial at 4°C for reanalysis if  necessary.

7.5   Analysis

      7.5.1 Transfer the  first  sample  vial  to an  autosampler  set  up to
inject  2.0  uL   portions  into  the  gas  chromatograph  for  analysis.
Alternately,  2 n(.  portions  of samples,  blanks  and  standards  may be
manually injected, using  the  solvent flush technique,  although  an auto
sampler is strongly recommended.

7.6   Determination of sample  volume

      7.6.1 For samples and field blanks, remove the cap from  the sample
container. Discard  the  remaining  sample/hexane mixture.   Shake  off the
remaining few drops using  short, brisk wrist movements. Reweigh the empty
container  with  original  cap and calculate  the net weight  of  sample by
difference to  the nearest 0.1 g.   This net weight is equivalent to the
volume of water extracted.

7.7   Calculations

      7.7.1 Identify EDB and DBCP  in  the sample chromatogram by comparing
the retention time of the  suspect peak to retention  times generated by the
calibration standards and the  check standard.

      7.7.2 Use the  calibration curve  or  calibration factor to directly
calculate the uncorrected  concentration  (C;) of each analyte  in  the sample
(e.g. calibration factor x response).

      7.7.3 Calculate the  sample volume  (Vs)  as  equal  to  the  net sample
weight:

      Vs (ml) = gross weight (grams)  -  bottle tare  (grams)

      7.7.4 Calculate the corrected sample concentration as:

      Concentration  (^9/L) =  C, x 35
      7.7.5 Report the results for the unknown samples in M9/L-  Round the

                             8011 - 6                         Revision 0
                                                               July 1992

-------
      results to the nearest  0.01  M9/L or two significant figures.


8.0   QUALITY CONTROL

      8.1   Each laboratory that uses this method is required to operate a formal
quality control  program.

            8.1.1 The laboratory  must make  an  initial  determination of  the
      method detection limits and demonstrate the ability to generate acceptable
      accuracy and precision with this method. This  is established as described
      in Section 8.2.

            8.1.2 In recognition of laboratory advances that are  occurring in
      chromatography, the laboratory is permitted certain options to improve the
      separations  or lower  the  cost of  measurements.     Each   time  such  a
      modification is made to the method, the analyst is required to repeat the
      procedure in Section  7.1  and 8.2.

            8.1.3 The laboratory must analyze a reagent  and calibration blank to
      demonstrate  that  interferences  from  the  analytical   system are  under
      control every twenty  samples or per  analytical batch,  whichever is more
      frequent.

            8.1.4 The laboratory must, on an ongoing basis,  demonstrate through
      the  analyses  of  QC  reference  samples  and  check standards  that  the
      operation of the  measurement system is in  control.  The frequency of the
      check  standard analyses   is  equivalent to  5% of all  samples  or  every
      analytical batch,  whichever is more frequent.   On a weekly basis,  the QC
      reference sample  must  be  run.

      8.2   To establ ish the abil ity to achieve low detection  1 imits and generate
acceptable  accuracy and precision,  the  analyst  must  perform  the  following
operations:

            8.2.1 Prepare  seven samples  each at  a  concentration  of 0.03 pg/L.

            8.2.2 Analyze  the  samples according  to the method  beginning in
      Section 7.0.

            8.2.3 Calculate  the average  concentration  (X)  in   ^g/L  and  the
      standard  deviation of  the concentrations  (s)  in ^g/L, for  each analyte
      using the seven results.   Then calculate the MDL  at 99% confidence level
      for seven replicates  as 3.143s.

            8.2.4 For each  analyte in  an  aqueous  matrix   sample,  X must be
      between 60% and 140% of  the true value.  Additionally, the MDL may not
      exceed  the  0.03  jig/L  spiked concentration.   If  both  analytes  meet the
      acceptance criteria,  the  system  performance  is acceptable and analysis of
      actual  samples can begin.   If either  analyte  fails to meet a criterion,
      repeat  the  test.  It  is  recommended that the laboratory  repeat  the MDL
      determination on  a regular basis.

      8.3   The laboratory must demonstrate on a frequency  equivalent to 5% of

                                   8011 - 7                         Revision 0
                                                                     July  1992

-------
the sample load or once per analytical batch, whichever is more frequent, that
the measurement  system is  in  control by  analyzing  a check  standard  of both
analytes at 0.25
            8.3.1 Prepare  a  check  standard   (0.25  jig/L)  by  diluting  the
      intermediate standard with water to 0.25  ng/L.

            8.3.2 Analyze the sample according to Section  7.0 and calculate the
      recovery for each analyte.  The recovery  must be between 60% and 140% of
      the expected value  for  aqueous  matrices.   For non-aqueous matrices, the
      U.S. EPA will set criteria after more  interlaboratory data are gathered.

            8.3.3 If  the  recovery  for  either  analyte  falls  outside  the
      designated range,  the  analyte fails the  acceptance  criteria.   A second
      calibration verification standard containing each analyte that failed must
      be  analyzed.  Repeated  failure,  however,  will confirm a general problem
      with the  measurement system.   If  this  occurs,  locate  and  correct the
      source of the problem and repeat the test.

      8.4   On a weekly  basis,  the  laboratory must demonstrate the ability to
analyze a QC reference sample.

            8.4.1 Prepare a QC reference sample at 0.10 yg/L by diluting the QC
      reference sample concentrate  (Section  5.9).

            8.4.2 For each analyte  in  an  aqueous matrix,  the recovery must be
      between 60% and  140% of  the expected value.  When either analyte fails the
      test, the analyst must repeat the test  only for that  analyte which failed
      to meet the criteria.   Repeated failure,  however, will confirm a genera\
      problem with the measurement  system or faulty samples and/or standards.
      If  this occurs, locate  and  correct  the source of the problem and repeat
      the test.  For non-aqueous matrices, the U.S.  EPA will set criteria after
      more interlaboratory data are gathered.

      8.5   Instrument  performance  -  Check the  performance  of the  entire
analytical system daily using data gathered from analyses  of blanks, standards,
and replicate samples.

            8.5.1 Peak tailing  significantly in  excess  of that  shown  in the
      chromatogram (Figure 1)  must be corrected.  Tailing  problems  are generally
      traceable to active  sites on  the GC column or to the detector operation.

            8.5.2 Check  the precision  between replicate  analyses.   A properly
      operating  system  should  perform  with  an   average relative  standard
      deviation  of less  than  10%.   Poor precision is generally  traceable to
      pneumatic leaks, especially at the  injection  port.


9.0   METHOD PERFORMANCE

      9.1   Method detection limits are presented in Table 1.  Single laboratory
accuracy  and precision at  several concentrations in tap water  are presented in
Table 2.


                                    8011 - 8                         Revision 0
                                                                     July  1992

-------
      9.2   In a preservation study extending over a  4 week period, the average
percent recoveries and  relative  standard  deviations  presented in Table 3 were
observed for organic-free reagent water (acidified), tap water and ground water.
The  results  for  acidified  and  non-acidified  samples  were  not  significantly
different.
10.0  REFERENCES

1.    Optimization of Liquid-Liquid Extraction Methods  for Analysis of Orqanics
      in Water. EPA-600/S4-83-052, 1984.

2.    Henderson, J.E.; Peyton, G.R.;  Glaze, W.H. Identification and Analysis of
      Organic Pollutants in Water; Keith, L.H., Ed; Ann Arbor Sci.: Ann Arbor,
      MI; 1976.

3.    Richard J.J.; Junk, G.A. Journal AWWA 1977, 69, 62.

4.    Budde, W.L.; Eichelberger, J.W. Organic Analyses Using Gas Chromatography-
      Mass Soectrometrv; Ann Arbor Science: Ann Arbor,  MI; 1978.

5.    Glaser, J.A.; et al.  Environmental  Science and Technology  1981, 15, 1426.

6.    Methods for the  Determination of  Organic Compounds in  Finished Drinking
      Water and Raw Source Water; U.S. Environmental Protection Agency. Office
      of  Research  and  Development.   Environmental  Monitoring  and  Support
      Laboratory.  ORD Publication Offices  of  Center for  Environmental Research
      Information:  Cincinnati,  OH 1986.
                                   8011 - 9                         Revision 0
                                                                     July 1992

-------
                                   TABLE 1.
               CHROMATOGRAPHIC CONDITIONS AND METHOD DETECTION
                 LIMITS  (MDL) FOR 1,2-OIBROMOETHANE (EDB) AND
                      l,2-DIBROMO-3-CHLOROPROPANE (DBCP)
Analyte
Retention Time.  Minutes
Column A    Column B      MDL (ng/l)
EDB
DBCP
 9.5
17.3
 8.9
15.0
0.01
0.01
Column A:  Durawax-DX 3
Column B:  DB-1
                                   TABLE 2.
                   SINGLE LABORATORY ACCURACY AND PRECISION
                         FOR  EDB  AND DBCP  IN TAP WATER



Analyte
EDB


DBCP



Number
of
Samples
7
7
7
7
7
7

Spike
Concentration
(MA)
0.03
0.24
50.0
0.03
0.24
50.0

Average
Recovery
(*)
114
98
95
90
102
94
Relative
Standard
Deviation
(%)
9.5
11.8
4.7
11.4
8.3
4.8
                                   8011  -  10
                                Revision 0
                                 July 1992

-------
                             TABLE 3.
                ACCURACY AND PRECISION AT 2.0
                    OVER A 4-WEEK STUDY PERIOD


Analyte
EOB




DBCP






Matrix1
RW-A
GW
GW-A
TW
TW-A
RW-A
GW
GW-A
TW
TW-A

Number
of Samples
16
15
16
16
16
16
16
16
16
16
Average
Accuracy
(% Recovery)
104
101
96
93
93
105
105
101
95
94
Relative
Std. Dev.
(%)
4.7
2.5
4.7
6.3
6.1
8.2
6.2
8.4
10.1
6.9
RW-A  =*     Organic-free reagent water at pH 2
GW    =     Ground water,  ambient pH
GW-A  =     Ground water at pH 2
TW    =     Tap water, ambient pH
TW-A  =     Tap water at pH 2
                             8011  -  11                         Revision 0
                                                               July  1992

-------
                    FIGURE 1.
SAMPLE CHROMATOGRAM FOR EXTRACT OF WATER SPIKED
        AT  0.114  M9/L WITH ED8 AND DBCP
                              COLUMN:   Fused silica  capillary
                              LIQUID PHASE:   Durawax-OX3
                              FILM THICKNESS:   0.25  \m
                              COLUMN DIMENSIONS:   30 M x 0.317
                                 ID
         10  IS  14   !•   It
                TIME  (MIN)
20   22   24
                    8011 - 12
                     Revision 0
                      July 1992

-------
                      METHOD  8011
1,2-DIBROMOETHANE AND  1,2-OIBROMO-3-CHLOROPROPANE
     BY MICROEXTRACTION AND GAS  CHROMATOGRAPHY
         Start
      7 2 Calibrate
      in* t r unent
        prepa re
      calibra tion
        curve
       7  2 Chick
      in* t r un*jn t
     performance
      7  3 Pr«par«
       lample*
  7  4.  1 Add
  VaCl ta
  sampl•*
  7  « 3 Add
 h»»«n« and
  • x t ract
  lamp 1•
                          7 4 4 Put
                           pair t of
                          tulracl  in
                            vial
 ? < 5 Sav,
remainder  of
 •Mtract for
  poiiibl*.
 raana i y>i*
 7  S Analyze
   by CC
7  6 Determine
  •ample
  volume
                   7  7 Calculate
                  concentration*
                                               Stop
                        8011  -  13
                                       Revision 0
                                        July  1992

-------
00
o
KS*

-------
                                 METHOD 80ISA

            NONHALOGENATED VOLATILE ORGANICS BY GAS CHROMATOGRAPHY
1.0   SCOPE AND APPLICATION

      1.1   Method  8015  is  used to  determine  the  concentration  of  various
nonhalogenated  volatile  organic  compounds.    The  following compounds  can  be
determined by this method:


                                                     Appropriate Technique
                                                                      Direct
Compound Name                        CAS No."       Purge-and-Trap     Injection
Diethyl ether
Ethanol
Methyl ethyl ketone (MEK)
Methyl isobutyl ketone (MIBK)
60-29-7
64-17-5
78-93-3
108-10-1
b
i
PP
PP
b
b
b
b
     a  Chemical Abstract Services Registry Number.
     b  Adequate response using this technique
     i  Inappropriate technique for this analyte
     pp Poor purging efficiency, resulting in high EQLs


2.0   SUMMARY OF METHOD

      2.1   Method 8015 provides gas chromatographic conditions for the detection
of certain nonhalogenated volatile organic compounds.  Samples may be introduced
into the GC using direct injection or purge-and-trap (Method 5030).  Ground water
samples must be analyzed by Method 5030.  A temperature program is used in the
gas chromatograph to separate the organic compounds.  Detection is achieved by
a flame ionization detector (FID).

      2.2   The method provides an  optional gas chromatographic column that may
be helpful in resolving the analytes from co-eluting non-target compounds and for
analyte confirmation.


3.0   INTERFERENCES

      3.1   Refer to Method 5030 and 8000.

      3.2   Samples  can  be  contaminated  by  diffusion  of volatile  organics
(particularly chlorofluorocarbons  and  methylene chloride) through  the  sample
container  septum  during shipment  and  storage.   A  trip blank  prepared  from
organic-free reagent water and carried  through sampling and subsequent storage
and handling can serve as a check on such contamination.


                                   8015A -  1                         Revision 1
                                                                     July 1992

-------
4.0   APPARATUS AND MATERIALS

      4.1   Gas chromatograph

            4.1.1 Gas  Chromatograph   -  Analytical  system  complete with  gas
      chromatograph suitable for on-column injections or purge-and-trap sample
      introduction and  all  required  accessories,  including  detectors,  column
      supplies, recorder,  gases,  and syringes.  A data system for measuring peak
      heights and/or peak areas is recommended.

            4.1.2 Columns

                  4.1.2.1     Column  1 - 8 ft x 0.1 in. ID  stainless steel or
            glass column  packed  with  1% SP-1000 on  Carbopack-B  60/80  mesh or
            equivalent.

                  4.1.2.2     Column  2 - 6 ft  x  0.1  in.  ID stainless steel or
            glass  column  packed  with   n-octane  on   Porasil-C  100/120  mesh
            (Durapak) or equivalent.

            4.1.3 Detector - Flame ionization (FID).

      4.2   Sample  introduction   apparatus  -  Refer   to  Method  5030 for  the
appropriate equipment for sample introduction purposes.

      4.3   Syringes - A 5 ml Luerlok glass hypodermic  and a 5 ml, gas-tight with
shutoff valve.

      4.4   Volumetric  flasks, Class  A  - Appropriate  sizes  with  ground glass
stoppers.

      4.5   Microsyringes - 10 and 25 nl with  a  0.006 in. ID needle (Hamilton
702N or equivalent) and a 100 nl.

      4.6   Analytical balance - 0.0001  g.


5.0   REAGENTS

      5.1   Reagent grade chemicals shall be used in all tests. Unless otherwise
indicated, it is intended that  all  reagents shall conform  to the specifications
of the Committee on Analytical  Reagents of the American Chemical Society, where
such specifications are available. Other grades  may  be used, provided it is first
ascertained that the  reagent is  of sufficiently  high purity to permit  its use
without lessening the accuracy of the determination.

      5.2   Organic-free reagent water -  All  references to water in this method
refer to organic-free reagent water,  as defined  in Chapter One.

      5.3   Methanol, CH3OH.   Pesticide quality or  equivalent.  Store away from
other solvents.

      5.4   Stock standards - Stock solutions may be prepared from pure standard
materials  or  purchased as  certified  solutions.    Prepare  stock  standards in

                                   8015A  -  2                         Revision 1
                                                                     July 1992

-------
methanol using assayed liquids.

            5.4.1 Place about 9.8 ml of methanol  in  a  10 ml tared, ground-glass
      stoppered volumetric flask.   Allow  the  flask  to stand,  unstoppered, for
      about 10 minutes or until all alcohol wetted surfaces have dried.  Weigh
      the flask to the nearest 0.0001 g.

            5.4.2 Using a 100 nl syringe,  immediately add two or more drops of
      assayed reference material to the  flask;  then reweigh.   The liquid must
      fall directly into the alcohol without contacting the neck of the flask.

            5.4.3 Reweigh, dilute to volume, stopper,  and then mix by inverting
      the flask several times.   Calculate the concentration  in milligrams per
      liter (mg/L) from the net gain in weight.  When compound purity is assayed
      to  be  96%  or  greater,  the  weight  may be used  without correction  to
      calculate the concentration of the stock standard.  Commercially prepared
      stock standards may be  used at any concentration if they are certified by
      the manufacturer or by an independent source.

            5.4.4 Transfer the  stock  standard  solution  into  a bottle  with  a
      Teflon lined screw-cap.  Store,  with minimal headspace,  at -10°C to  -20°C
      and protect from light.

            5.4.5 Standards  must be  replaced after  6 months,  or sooner  if
      comparison with check standards indicates a problem.

      5.5   Secondary dilution standards - Using  stock standard solutions, pre-
pare  in  methanol  secondary  dilution  standards, as  needed,  that  contain the
compounds of interest, either singly or mixed  together.  The secondary dilution
standards should be prepared at concentrations  such that the aqueous calibration
standards  prepared   in  Section  5.5 will  bracket   the  working  range  of the
analytical system.  Secondary dilution standards should be stored with minimal
headspace for volatiles  and should be checked frequently for  signs of degradation
or evaporation, especially just  prior  to  preparing  calibration standards from
them.

      5.6   Calibration standards - Calibration standards at a minimum of five
concentrations are prepared  in water  from the secondary  dilution  of the stock
standards.  One  of  the concentrations should  be at. a  concentration near, but
above,  the  method   detection  limit.    The  remaining  concentrations  should
correspond to  the expected range  of  concentrations found  in  real  samples  or
should define the working range  of the GC. Each standard  should contain each
analyte for detection by this method (e.g. some or all of the compounds listed
in Section 1.1 may be included).   In order*to  prepare accurate aqueous standard
solutions, the following precautions must  be  observed:

            5.6.1 Do  not  inject  more  than  20 ^L of  alcoholic  standards into
      100 ml of water.

            5.6.2 Use  a  25  ^L  Hamilton  702N  microsyringe  or  equivalent
      (variations  in needle  geometry will  adversely affect  the  ability  to
      deliver reproducible volumes of methanolic standards into water).

            5.6.3 Rapidly  inject  the  alcoholic  standard  into  the  filled

                                   8015A - 3                        Revision  1
                                                                     July 1992

-------
      volumetric flask.  Remove the needle as fast  as possible after injection.

            5.6.4 Mix aqueous standards by inverting the flask three times only.

            5.6.5 Fill the sample syringe  from the standard solution contained
      in the expanded  area of  the  flask (do not use any solution contained in
      the neck of the  flask).

            5.6.6 Never  use  pipets to  dilute or transfer  samples  or  aqueous
      standards.

            5.6.7 Aqueous standards are not stable and should be discarded after
      1 hour, unless properly sealed and  stored.  The aqueous standards can be
      stored up to 24  hours,  if held in sealed vials with zero headspace.

      5.7   Internal standards (if internal  standard calibration is used) - To
use this approach, the analyst must select  one or more internal  standards that
are similar in  analytical behavior  to  the compounds of interest.   The  analyst
must further demonstrate  that  the  measurement of the internal  standard is not
affected by method or  matrix  interferences.  Because of these  limitations,  no
internal standard can  be suggested that is  applicable to all samples.

            5.7.1 Prepare  calibration  standards   at   a  minimum  of   five
      concentrations for each parameter of interest as described  in Section 5.6.

            5.7.2 Prepare a  spiking  solution containing each of  the  internal
      standards using  the procedures described in Sections 5.4 and 5,5.  It is
      recommended  that  the   secondary dilution  standard  be   prepared  at  a
      concentration  of  15  ng//xL  of   each  internal  standard  compound.    The
      addition  of  10 pi of  this  standard to 5.0 ml of  sample  or calibration
      standard would be equivalent to  30
            5.7.3 Analyze each calibration  standard  according  to Section 7.0,
      adding  10 pi.  of  internal  standard  spiking  solution  directly to  the
      syringe.

      5.8   Surrogate standards - The analyst should monitor both the performance
of the analytical  system and the effectiveness of the method in dealing with each
sample matrix by spiking each sample,  standard,  and water  blank with one or two
surrogate compounds  recommended to  encompass the  range of temperature program
used in this method.   From stock standard solutions prepared as in Section 5.4,
add a volume to give 750 ng of each surrogate to 45 ml of water contained in a
50  ml  volumetric  flask,  mix, and  dilute  to volume  for a concentration  of
15 ng//iL.  Add  10 nl of this  surrogate spiking  solution directly into the 5 ml
syringe with  every sample and reference standard  analyzed.    If the  internal
standard calibration procedure is  used,  the  surrogate compounds  may  be  added
directly to the internal standard spiking solution (Section 5.7.2).


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See the  introductory  material to this Chapter,  Organic Analytes,
Section 4.1.
                                   8015A -  4                        Revision 1
                                                                     July 1992

-------
7.0   PROCEDURE

      7.1   Volatile compounds are introduced into  the gas chromatograph either
by direct injection or purge-and-trap  (Method  5030).   Method 5030 may be used
directly on ground  water samples or low-concentration  contaminated soils and
sediments.   For high-concentration soils or sediments, methanolic extraction, as
described in Method  5030,  may be necessary  prior  to  purge-and-trap analysis.
Method 5030 also  provides guidance on the analysis  of aqueous miscible and non-
aqueous miscible  liquid wastes (see Section 7.4.1.1).

      7.2   Chromatographic conditions (Recommended)

            7.2.1 Column 1

            Carrier gas (Helium) flow rate:     40 mL/min
            Temperature program:
                  Initial temperature:           45°C,  hold  for 3  minutes
                  Program:                      45°C to  220°C at 8°C/nnn
                  Final temperature:            220°C,  hold for 15 minutes.

            7.2.2 Column 2

            Carrier gas (Helium) flow rate:     40 mL/min
            Temperature program:
                  Initial temperature:           50°C,  hold  for 3  minutes
                  Program:                      50°C to  170°C  at 6°C/min
                  Final temperature:            170°C,  hold for 4 minutes.

      7.3   Calibration - Refer to Method 8000 for proper  calibration techniques.

            7.3.1 Calibration must take place using the same sample introduction
      method that will be used to analyze actual samples (see Section 7.4.1).

            7.3.2 The  procedure  for internal  or  external  calibration may be
      used.  Refer to  Method 8000 for a description  of each of these procedures.

      7.4   Gas chromatographic analysis

            7.4.1 Introduce volatile compounds  into the  gas chromatograph using
      either Method 5030  (purge-and-trap method) or the direct injection method.
      If the  internal standard  calibration  technique  is  used,  add  10 jiL of
      internal standard to the sample prior to purging.

                  7.4.1.1     Direct injection  -  In very limited applications
            (e.g. aqueous process wastes), direct injection of the sample into
            the GC  system with  a 10 pi  syringe may be  appropriate.   One such
            application is for verification of the alcohol content  of an aqueous
            sample prior to determining if the sample is  ignitable  (Methods 1010
            or 1020).  In this  case,  it  is  suggested  that  direct injection be
            used. The detection limit  is  very high  (approximately 10,000 M9/L);
            therefore, it  is  only permitted when  concentrations  in excess of
            10,000 A*9/L are expected or for water-soluble compounds  that do not
            purge.  The system must be calibrated by direct injection (bypassing
            the purge-and-trap device).

                                   8015A - 5                         Revision 1
                                                                     July 1992

-------
                  Non-aqueous  miscible wastes may  also be analyzed  by  direct
            injection if the  concentration  of target, analytes  in the  sample
            falls within the calibration range.   If dilution of the  sample  is
            necessary,  follow the guidance  for  High Concentration samples  in
            Method 5030, Section  7.3.3.2.

            7.4.2 Method 8000  provides instructions on  the analysis  sequence,
      appropriate dilutions,  establishing  daily retention  time windows,  and
      identification criteria.  Include a mid-concentration standard after each
      group of 10 samples in the  analysis  sequence.*

            7.4.3 Record the sample volume purged or injected and the  resulting
      peak sizes (in area units or peak heights).

            7.4.4 Calculation  of  concentration is  covered in  Method 8000.

            7.4.5 If analytical interferences  are suspected, or for the purpose
      of confirmation,  analysis using the  second GC column is recommended.

            7.4.6 If the response  for a peak is off-scale,  prepare a dilution of
      the sample with water.  The  dilution  must be performed on a second aliquot
      of the sample which has  been properly sealed  and stored prior to use.


8.0   QUALITY CONTROL

      8.1   Refer to Chapter  One  for specific quality  control  procedures  and
Method 8000 for gas chromatographic procedures.  Quality control to ensure the
proper operation of the purge-and-trap device is covered in Method 5030.

      8.2   Quality control  required to validate the GC system  operation is found
in Method 8000, Section 8.6.

      8.3   Calculate surrogate standard recovery  on  all  samples,  blanks,  and
spikes.  Determine if recovery  is within limits (limits  established by performing
QC procedure outlined in Method 8000, Section 8.10).

            8.3.1 If recovery is  not within limits, the following is  required:

            •     Check  to  be sure  that there are  no  errors  in calculations,
                  surrogate solutions,  and internal  standards.    Also,  check
                  instrument performance.

            •     Recalculate the data and/or reanalyze the  extract  if  any of
                  the above checks reveal  a problem.

            •     Re-extract and  re-analyze the  sample  if none of the above are
                  a problem or flag the data as  "estimated concentration".


9.0   METHOD PERFORMANCE

      9.1   The accuracy and precision obtained will be determined by the sample
matrix,  sample  introduction technique, and calibration  procedures used.

                                   8015A -  6                        Revision  1
                                                                     July 1992

-------
      9.2   Specific
becomes available.
method  performance  information will  be provided  as  it
10.0  REFERENCES

1.    Bellar,  T.A.,  and  J.J.  Lichtenberg,  Determining  Volatile  Organics  at
      Microgram-per-Liter Levels  by  Gas Chromatography,  J. Amer.  Water Works
      Assoc., 66(121. pp. 739-744 (1974).

2.    Bellar, T.A., and J.J. Lichtenberg, Semi-Automated  Headspace Analysis of
      Drinking  Waters  and  Industrial  Waters for  Purgeable Volatile  Organic
      Compounds, in Van  Hall, ed., Measurement  of  Organic  Pollutants in Water
      and Wastewater, ASTM STP 686,  pp. 108-129, 1979.

3.    Development and Application of  Test Procedures for Specific Organic Toxic
      Substances in Wastewaters:  Category 11 - Purgeables and  Category  12 -
      Acrolein,  Acrylonitrile,   and  Dichlorodifluoromethane,   Report for  EPA
      Contract 68-03-2635 (in preparation).
                                   8015A -  7
                                               Revision  1
                                                July  1992

-------
                         METHOD 8015A
NONHALOGENATED VOLATILE ORGANICS BY GAS CHROMATOGRAPHY
                            voluae purged
       chroma, tog r a phi c
                           7 1 5 Calculate
                           concentration*
                              7 4. 4 Are
                             analytical
                            interference*
                             iuapeeled?
746 Analyie
sample
  aeeonot CC
   eeluan
•emple into CC
      purge-and-trap
                            7 4  7 la peak
                            reaponae off
                               tea!.'
                            8015A - 8
                   Revision  1
                    July 1992

-------
                                 METHOD 8015B

                   NONHALOGENATED ORGANICS USING GC/FID
1.0   SCOPE AND APPLICATION

     1.1  Method 8015 is used to determine the concentration of various nonhalogenated volatile
organic compounds and semivolatile organic compounds by gas chromatography. The following
compounds can be determined quantitatively by this method:
Appropriate Techniaue
Compound Name
Acetone
Acetonitrile
Acrolein
Acrylonitrile
Allyl alcohol
1-Butanol (n-Butyl alcohol)
t-Butyl alcohol
2-Chloroacrylonitrile (I.S.)
Crotonaldehyde
Diethyl ether
1 ,4-Dioxane
Ethanol
Ethyl acetate
Ethylene glycol
Ethylene oxide
Hexafluoro-2-propanol (I.S.)
Hexafluoro-2-methyl-
2-propanol (I.S.)
Isobutyl alcohol
Isopropyl alcohol
Methanol
Methyl ethyl ketone (MEK)
Methyl isobutyl ketone (MIBK)
N-Nitroso-di-n-butylamine
Paraldehyde
2-Pentanone
2-Picoline
1-Propanol
Propionitrile
CAS No.8
67-64-1
75-05-8
107-02-8
107-13-1
107-18-6
71-36-3
75-65-0
920-37-6
123-73-9
60-29-7
123-91-1
64-17-5
141-78-6
107-21-1
75-21-8
920-66-1

515-14-6
78-83-1
67-63-0
67-56-1
78-93-3
108-10-1
924-16-3
123-63-7
107-87-9
109-06-8
71-23-8
107-12-0
Purge-and-
Trap
PP
PP
PP
PP
ht
ht
PP
NA
PP
b
PP
I
I
I
I
NA

NA
PP
PP
I
PP
PP
PP
PP
PP
PP
PP
ht
Direct
Injection
b,d
b,d
b.d
b,d
b,d
b,d
b,d
d
b,d
b
b,d
b,d
b,d
b
b,d
d

d
b,d
b,d
b,d
b,d
b,d
b,d
b,d
b.d
b,d
b.d
d
Solvent
Extraction







h














JA







NA

NA
I
I
I
I
I
b
I
I
I
I
I
CD-ROM
8015B-1
    Revision 2
December 1996

-------
                                             	Appropriate Technique	
Compound Name                             Purge-and-    Direct     Solvent
                              CAS No.a          Trap      Injection    Extraction
Pyridine                       110-86-1           I           b,d              b
o-Toluidine                     95-53-4           I           b,d              b
  8     Chemical Abstract Services Registry Number.
  b    Adequate response using this technique
  d    Amenable to concentration by azeotropic distillation (Method 5031)
 ht    Method analyte only when purged at 80°C
  I     Inappropriate technique for this analyte
 pp    Poor purging efficiency, resulting in high EQLs
 NA   Not available
I.S.    Internal standard appropriate for Method 5031

     1.2  This method may also be applicable to the analysis of petroleum hydrocarbons, including
gasoline range organics (GROs) and diesel range organics (DROs). GROs correspond to the range
of alkanes from  C6 to C10and covering a boiling  point range of approximately 60°C - 170°C
(Reference 6). DROs correspond to the range of alkanes from C10 to C28 and covering a boiling point
range of approximately 170°C - 430°C (Reference 6).  The identification of specific fuel types may
be complicated by environmental processes such as evaporation, biodegradation, or when more
than one fuel type is present. Methods from other sources may be more appropriate for GROs and
DROs  since these hydrocarbons are not regulated under RCRA.  Consult State and local regulatory
authorities for specific requirements.

     1.3  This method is restricted for use by, or under the supervision of, analysts experienced
in the use of gas chromatographs and skilled in the interpretation of gas chromatograms.  In addition,
if this method is used for the analysis of petroleum hydrocarbons, it is limited to analysts experienced
in the interpretation of hydrocarbon  data.  Each analyst must demonstrate the ability to generate
acceptable results with this method.

     1.4  The method can also be used  as a screening tool (for both volatile and semivolatile
organics) to obtain semiquantitative data for the prevention of sample overload during quantitative
analysis on  a GC/MS system. This may  be accomplished  using an  automated (Method 5021)
headspace method or by direct injection if a solvent extraction method has been utilized for sample
preparation.  Single point calibration would be acceptable in this situation. Performance data are
not provided for screening.


2.0  SUMMARY OF METHOD

     2.1  Method 8015  provides gas chromatographic conditions for the detection of  certain
nonhalogenated volatile and semivolatile organic compounds.

          2.1.1   Samples may be introduced into the GC:

                following solvent  extraction (Methods 3510, 3520, 3540, 3541, 3545, 3550, or
                3560)

CD-ROM                               8015B - 2                               Revision 2
                                                                         December 1996

-------
                 by direct injection (aqueous samples) including the concentration of analytes by
                 azeotropic distillation (Method 5031)

                 by purge-and-trap (Methods 5030 or 5035), or

                 by vacuum distillation (Method 5032)

          2.1.2   Ground or surface water samples must generally be analyzed in conjunction with
     Methods 5030, 5031, 5032, 3510, 3520, or other appropriate preparatory methods to obtain
     the necessary quantitation limits. Method 3535 (solid-phase extraction) may also be applicable
     to the target analytes, but has not yet been validated by EPA in conjunction with Method 8015.

          2.1.3   Diesel range  organics (DROs) may  be prepared by an  appropriate  solvent
     extraction method.

          2.1.4   Gasoline range organics (GROs) may be introduced into the GC/FID by purge-
     and-trap, automated headspace, vacuum distillation, or other appropriate technique.

     2.2  An appropriate column and temperature program is used in the gas chromatograph to
separate the organic compounds. Detection  is achieved by a flame ionization detector (FID).

     2.3  The  method allows the use of  packed or capillary columns for the analysis and
confirmation of the non-halogenated individual analytes.  Columns and conditions listed have been
demonstrated to provide separation of those target analytes.  Analysts may change these conditions
as long as they demonstrate adequate performance.

     2.4  Fused silica capillary columns are necessary for the analysis of petroleum hydrocarbons.


3.0  INTERFERENCES

     3.1  When analyzing for volatile organics, samples can be contaminated by diffusion of volatile
organics (particularly chlorofluorocarbons and methylene chloride) through the sample container
septum during shipment and storage.  A trip blank prepared from organic-free reagent water and
carried through  sampling and subsequent storage and handling must serve as a check on such
contamination.

     3.2  Contamination  by   carryover   can   occur  whenever  high-concentration  and
low-concentration samples are analyzed in sequence. To reduce the potential for carryover, the
sample syringe or purging device must be rinsed out between samples with an appropriate solvent.
Whenever an unusually concentrated sample is  encountered, it should be followed by injection of
a solvent blank to check for cross contamination.

          3.2.1   Clean purging vessels with a detergent solution, rinse with distilled water, and
     then dry in a 105°C oven between analyses. Clean syringes or autosamplers by flushing all
     surfaces that contact samples using appropriate solvents.

          3.2.2   All glassware must be scrupulously cleaned.  Clean all glassware as soon as
     possible after use by rinsing with the last solvent used. This should be followed by detergent
     washing with hot water, and rinses with tap water and organic-free reagent water. Drain the
     glassware and dry in an oven at 130°C for several hours or rinse with methanol and drain.
     Store dry glassware in a clean environment.


CD-ROM                               8015B-3                              Revision 2
                                                                         December 1996

-------
     3.3  The flame ionization detector (FID) is a non-selective detector.  There is a potential for
many non-target compounds present in samples to interfere with this analysis.


4.0  APPARATUS AND MATERIALS

     4.1  Gas chromatograph

          4.1.1   Gas  Chromatograph - Analytical system complete with gas chromatograph
     suitable for solvent  injections or purge-and-trap sample  introduction  and  all required
     accessories, including detectors, column supplies, recorder, gases, and syringes.  A data
     system for measuring peak heights and/or peak areas is recommended.

          4.1.2   Recommended GC Columns

                 4.1.2.1  Column 1 - 8 ft x 0.1 in. ID stainless steel or glass column packed with
          1% SP-1000 on Carbopack-B 60/80 mesh or equivalent.

                 4.1.2.2  Column 2 - 6 ft x 0.1 in. ID stainless steel or glass column packed with
          n-octane on  Porasil-C 100/120 mesh (Durapak) or equivalent.

                 4.1.2.3  Column 3 - 30 m x 0.53 mm ID fused silica capillary column bonded with
          DB-Wax (or equivalent), 1-um film thickness.

                 4.1.2.4  Column 4 - 30 m x 0.53 mm ID fused silica capillary column chemically
          bonded with 5% methyl silicone (DB-5, SPB-5, RTx, or equivalent), 1.5-um film thickness.

                         4.1.2.4.1    Capillary columns are needed for petroleum hydrocarbon
                 analyses. Laboratories may use other capillary  columns (e.g. 0.25-0.32 mm ID
                 capillary  columns)  if  they  document  method   performance  data (e.g.
                 chromatographic resolution and  MDLs) if appropriate for the intended use of the
                 data.

                         4.1.2.4.2    Wide-bore columns  should  be  installed  in 1/4-inch
                 injectors, with deactivated liners designed specifically for use with these columns.

          4.1.3   Detector - Flame ionization (FID)

     4.2  Sample introduction and preparation apparatus

          4.2.1   Refer to  the 5000 series  sample preparation methods for the  appropriate
     apparatus.

          4.2.2   Samples may also be introduced into the GC via injection of solvent extracts or
     direct injection of aqueous samples.

     4.3  Syringes

          4.3.1   A 5-mL Luer-Lok glass hypodermic and a 5-mL gas-tight syringe with  shutoff
     valve for volatile analytes.
CD-ROM                               8015B-4                             Revision 2
                                                                        December 1996

-------
          4.3.2  Microsyringes -  10- and 25-ul_ with a 0.006 in. ID needle (Hamilton 702N or
     equivalent) and 100-uL

     4.4  Volumetric flasks, Class A - Appropriate sizes with ground glass stoppers.

     4.5  Analytical balance - 0 -160 g capacity, capable of measuring differences of 0.0001 g.


5.0  REAGENTS

     5.1  Reagent grade chemicals shall be used whenever possible. Unless otherwise indicated,
it is intended that all reagents shall conform to the specifications of the Committee on Analytical
Reagents of the American Chemical Society, where such specifications are available. Other grades
may be used, provided it is first ascertained that the reagent is of sufficiently high purity to permit its
use without lessening the accuracy of the determination.

     5.2  Organic-free reagent water - All references to water in this method refer to organic-free
reagent water, as defined in Chapter One.

     5.3  Methanol, CH3OH. Pesticide quality or equivalent. Store away from other solvents.

     5.4  Fuels, e.g., gasoline or diesel.  Purchase from a commercial source.  Low boiling
components in fuel evaporate quickly.  If available, obtain fuel from the leaking tank on  site.

     5.5  Alkane  standard.   A standard  containing a homologous  series of n-alkanes for
establishing retention times (e.g., C10-C32 for diesel).

     5.6  Stock standards - Stock solutions may be prepared from pure standard materials or
purchased as certified solutions.  When methanol is a target analyte or when using  azeotropic
distillation for sample preparation, standards should not be prepared in methanol. Standards must
be replaced after 6 months or sooner, if comparison with check standards indicates a problem.

     5.7  Secondary dilution standards - Using stock standard solutions, prepare secondary dilution
standards, as needed, that contain the  compounds of interest, either singly or mixed together. The
secondary dilution standards should be prepared at concentrations such that the aqueous  calibration
standards prepared in Sec. 5.8 will bracket the working range of the analytical system. Secondary
dilution standards should be stored with minimal headspace for volatiles and should be checked
frequently for signs of degradation or evaporation, especially just prior to preparing  calibration
standards from  them.

     5.8  Calibration standards - Calibration standards at a minimum of five different  concentrations
are prepared in water (purge-and-trap or direct  injection) or in methylene chloride (solvent injection)
from the secondary dilution of the stock standards. One of the standards should  be at or below the
concentration equivalent to the appropriate quantitation limit for the  project.  The  remaining
concentrations should correspond to  the expected range of concentrations found in real samples or
should define the working range of the GC. Each standard should contain each analyte for detection
by this method  (e.g., some or all of the compounds listed in Sec. 1.1 may be included).  Volatile
organic standards are prepared in organic-free reagent water.  In order to prepare accurate aqueous
standard solutions, the following precautions must be observed:

          5.8.1  Do not inject  more than 20 uL of methanolic standards into 100 mL of water.
CD-ROM                                8015B - 5                              Revision 2
                                                                          December 1996

-------
          5.8.2  Use a 25-pL Hamilton 702N microsyringe or equivalent (variations in needle
     geometry will adversely  affect the ability  to deliver reproducible volumes of methanolic
     standards into water).

          5.8.3  Rapidly inject the primary standard into the filled volumetric flask.  Remove the
     needle as fast as possible after injection.

          5.8.4  Mix diluted standards by inverting the flask three times only.

          5.8.5  Fill the sample syringe from the  standard solution contained in the expanded area
     of the flask (do not use any solution contained in the neck of the flask).

          5.8.6  Never use pipets to dilute or transfer samples or aqueous standards when diluting
     volatile organic standards.

          5.8.7  Aqueous standards  used for purge-and-trap analyses (Method 5030) are not
     stable and should be discarded after 1 hour,  unless held in sealed vials with zero headspace.
     If so stored, they may be held for up to 24 hours.  Aqueous  standards used for azeotropic
     distillation (Method 5031) may be stored  for  up to a month in polytetrafluoroethylene (PTFE)-
     sealed screw-cap bottles with minimal headspace, at 4°C, and protected from light.

     5.9  Internal standards (if internal standard calibration is used) - To use this approach, the
analyst must select one or more internal standards that are similar in analytical  behavior to the
compounds of interest.  The analyst must further demonstrate that the measurement of the internal
standard is not affected by method or matrix interferences.  Because of these limitations, no internal
standard can be suggested that is applicable to all samples. The following internal standards are
recommended when  preparing  samples  by  azeotropic  distillation:    2-chloroacrytonitrile,
hexafluoro-2-propanol and hexafluoro-2-methyl-2-propanol.

     5.10 Surrogate  standards -  Whenever possible,  the analyst should monitor  both the
performance of the analytical system and  the effectiveness of the method in dealing with each
sample matrix by spiking each sample, standard,  and blank with one or two surrogate compounds
which are not affected by method interferences.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     See the introductory material to this chapter, Organic Analytes, Sec. 4.1.


7.0  PROCEDURE

     7.1   Introduction/preparation methods

     Various  alternate methods are provided for  sample introduction.  All internal standards,
surrogates, and matrix spikes (when applicable) must be added to samples before introduction into
the GC/FID system. Follow the introduction method on when to add standards.
CD-ROM                               8015B-6                              Revision 2
                                                                         December 1996

-------
          7.1.1   Direct injection - This involves direct syringe injection into the GC injection port.

                 7.1.1.1   Volatile organics (includes gasoline range organics [GROs])

                 This may involve  injection of an aqueous sample containing  a very high
          concentration  of analytes; injection  of aqueous concentrates  from Method 5031
          (azeotropic distillation for nonpurgeable volatile organics); and injection of an organic
          solvent waste.  Direct injection of aqueous samples (non-concentrated) has very limited
          applications.   It is  only permitted for the determination of volatiles at the  toxicity
          characteristic (TC) regulatory limits or at concentrations in excess of 10,000 ug/L. It may
          also be used in conjunction with the test for ignitability in aqueous samples (along with
          Methods 1010 and 1020) to determine if alcohol is present at > 24%.

                 7.1.1.2  Semivolatile organics (includes diesel range organics [DROs])

                 This may involve syringe injection of extracts of aqueous samples prepared by
          Methods 3510 or 3520 or extracts of soil/solids prepared by Methods 3540, 3541, 3545,
          3550 or 3560.

                 WARNING:  Ultrasonic extraction (Method 3550) is  not as rigorous a method as
                             the other extraction methods for soil/solids. This means it is very
                             critical that the method be followed explicitly to achieve extraction
                             efficiency which approaches that of Soxhlet  extraction.  Consult
                             Method 3550 for information on the critical aspects of this extraction
                             procedure.

          7.1.2   Purge and trap - this includes purge and trap for aqueous samples (Method 5030)
     and purge and trap for solid samples (Method 5035).  Method 5035 also  provides techniques
     for extraction of solid  and oily waste samples by methanol (and other water miscible solvents)
     with subsequent purge and trap from an aqueous matrix using Method 5030. Normally purge
     and trap for aqueous samples is performed at ambient temperatures while soil/solid samples
     utilize a 40°C purge to improve extraction efficiency.  Occasionally, there may  be a need to
     perform a heated purge for aqueous samples to lower detection limits; however, a 25-mL
     sample should provide the sensitivity needed in most situations.

          7.1.3   Vacuum distillation - this  is a device for the introduction of volatile organics from
     aqueous, solid or tissue samples (Method 5032) into the GC/FID system.

          7.1.4   Automated static headspace - this is a device for the  introduction of volatile
     organics from solid samples (Method 5021) into the GC/FID system.

     7.2  Chromatographic conditions (recommended)

          7.2.1   Column 1

          Carrier gas (Helium) flow rate:    40 mL/min
          Temperature program:
                 Initial temperature:        45°C, hold for 3 minutes
                 Program:                45°C to 220°C at 8°C/min
                 Final temperature:        220°C, hold for 15 minutes.
CD-ROM                               8015B-7                              Revision 2
                                                                         December 1996

-------
          7.2.2  Column 2

          Carrier gas (Helium) flow rate:
          Temperature program:
                Initial temperature:
                Program:
                Final temperature:

          7.2.3  Column 3

          Carrier gas (Helium) flow rate:
          Temperature program:
                Initial temperature:
                Program:
                Final temperature:

          7.2.4  Column 4 (DROs)

          Carrier gas (Helium) flow rate:
          Makeup gas (Helium) flow rate:
          Injector temperature:
          Detector temperature:
          Temperature program:
                Initial temperature:
                Program:
                Final temperature:

          7.2.5  Column 4 (GROs)

          Carrier gas (Helium) flow rate:
          Makeup gas (Helium) flow rate:
          Injector temperature:
          Detector temperature:
          Temperature program:
                Initial temperature:
                Program:
                Final temperature:
                Final hold:

     7.3  Initial calibration
 40 mL/min

 50°C, hold for 3 minutes
 50°C to 170°C at 6°C/min
 170°C, hold for 4 minutes.
 15 mL/min

 45°C, hold for 4 minutes
 45°C to 220°C at 12°C/min
 220°C, hold for 3 minutes.
 5-7 mL/minute
 30 mL/min
 200°C
 340°C

 45°C, hold 3 minute
 45°Cto275°Cat12°C/min
 275°C, hold 12 min
 5-7 mL/minute
 30 mL/min
 200°C
 340°C

 45°C, hold 1 minute
 45°Cto100°Cat5°C/min
 100°C to 275°C, at 8°C/min
 5 min
          7.3.1  Set up the sample introduction system as outlined in the method of choice (see
     Sec. 7.1).  A different calibration curve is necessary for each sample introduction  mode
     because of the differences in conditions and equipment. Establish chromatographic operating
     parameters that provide instrument performance equivalent to that documented in this method.
     Prepare calibration standards using the procedures described above (Sec. 5.8).  The external
     standard technique  is described below.  Analysts wishing  to use the internal standard
     technique are referred  to  Method  8000.   Recommended internal  standards  for the
     non-purgeable volatiles include hexafluoro-2-propanol, hexafluoro-2-methyl-2-propanol, and
     2-chloroacrylonitrile.
CD-ROM
8015B-8
     Revision 2
December 1996

-------
          7.3.2  External standard calibration procedure for single component analytes

                 7.3.2.1   For each analyte and surrogate of interest, prepare calibration standards
          at a minimum of five different concentrations by adding volumes of one or more stock
          standards to a volumetric flask and diluting to volume with an appropriate solvent. One
          of the external standards should be at a concentration at or below the quantitation limit
          necessary for the project (based on the concentration in the final volume specified in the
          preparation method, with no dilutions).  The other concentrations should correspond to
          the expected range of concentrations found in real  samples or should define the working
          range of the detector.

                 7.3.2.2   Introduce each calibration standard using the technique that will be used
          to introduce the actual samples into the gas chromatograph.  Tabulate peak height or
          area responses against the mass injected.  Calculate the calibration factor (CF) for each
          single component analyte as described in Method 8000.

          7.3.3  External standard calibration procedure for DROs and  GROs

          The calibration of DROs and GROs is markedly different from that for single component
     analytes. In particular, the response used for calibration must represent the entire area of the
     chromatogram within the retention time range for the fuel type (DROs or GROs), including the
     unresolved complex mixture  that lies below  the  individual  peaks.   See Sec. 7.7.2 for
     information on calculating this area.

                 7.3.3.1   For each fuel type, prepare calibration standards at a minimum of five
          different concentrations by  adding volumes  of  one or  more stock  standards to a
          volumetric flask and diluting to volume with an appropriate solvent.  One of the external
          standards should be at a  concentration at or below the quantitation limit necessary for
          the project (based on the concentration in the final volume specified in the preparation
          method, with no dilutions). The other concentrations should correspond to the expected
          range of concentrations found in real samples or should define the working  range of the
          detector.

                 NOTE:   Whenever possible,  the calibration should be performed using the
                          specific fuel that is contaminating the site (e.g., a sample of the fuel
                          remaining in the tank suspected of leaking). Where such samples are
                          not available  or not  known,  use recently purchased commercially-
                          available fuel.  A qualitative screening injection and GC run may be
                          performed to identify unknown fuels.

                 7.3.3.2   Introduce each calibration standard using the technique that will be used
          to introduce the actual samples into the gas chromatograph.  Determine the area of the
          response as described in Sec. 7.7.2. Calculate the calibration  factor (CF) for each fuel
          type as shown below:

                ~ ...  ..    ,_   .      Total Area within Retention Time Range
                Calibrator! Factor  =       Mgss jnjected (jp nanograms)


     7.3.4       Calibration linearity

          The linearity of  the calibration  must  be assessed.  This applies  to both the single
     component analytes and the fuel types.
CD-ROM                                8015B - 9                               Revision 2
                                                                          December 1996

-------
                 7.3.4.1   If the percent relative standard deviation (%RSD) of the calibration
          factor is less than  20% over the working range, linearity through the origin can be
          assumed, and the average calibration factor can be used in place of a calibration curve.

                 7.3.4.2  If the % RSD is more than 20% over the working range, linearity through
          the origin cannot be assumed. See Method 8000 for other calibration options that may
          be employed.

     7.4  Retention time windows

          Single component target analytes (see Sec. 1.1) are identified on the basis of retention
     time windows. GROs and DROs are distinguished on the basis of the ranges of retention
     times for characteristic components in each type of fuel.

          7.4.1   Before establishing retention time windows, make sure that the  chromatographic
     system is functioning reliably  and that the operating parameters have been optimized for the
     target analytes and surrogates in the sample matrix to be analyzed. Establish  the retention
     time windows for single component target analytes using the procedure described in Sec. 7.0
     of Method  8000.

          7.4.2  The retention time range for GROs is defined during initial calibration.  Two
     specific gasoline components are  used to establish the range, 2-methylpentane and 1,2,4-
     trimethylbenzene. Use the procedure described in Sec. 7.0 of Method 8000 to  establish the
     retention time windows for these two components. The retention time range is then calculated
     based on the lower limit of the RT window for the first eluting component and the upper limit
     of the RT window for the last eluting component.

          7.4.3  The retention time range for DROs is defined during initial calibration. The range
     is established from the  retention times of the C10 and C 2a alkanes.   Use the procedure
     described  in Sec. 7.0 of Method 8000 to establish the retention time windows for these two
     components. The retention time range is then  calculated based on the  lower limit of the RT
     window for the first eluting component and the upper limit of the RT window for the last eluting
     component.

     7.5  Calibration verification

          7.5.1   The working calibration curve, and  retention  times  must be  verified at the
     beginning of each 12-hour work shift as a minimum requirement. Verification is accomplished
     by the measurement of one or more calibration standards (normally mid-concentration) that
     contain all of the target analytes and surrogates when individual target analytes are being
     analyzed.  Verification is accomplished by the measurement of the fuel standard and the
     hydrocarbon retention time standard when petroleum hydrocarbons are being  analyzed.
     Additional analyses of the verification standard(s) throughout a 12-hour shift are strongly
     recommended, especially for samples that contain visible concentrations of oily material. See
     Sec. 7.0 "calibration verification" of Method 8000 for more detailed  information.

          7.5.2  Calculate the % difference  as detailed in Sec.  7.0  of Method 8000.   If the
     response for any analyte is within ±15% of the response obtained during the initial calibration,
     then the initial calibration is considered still valid, and analyst may continue to use the mean
     CF or  RF values from the initial  calibration to quantitate  sample results.  For  analyses
     employing azeotropic distillation as the sample  introduction technique, the % difference may
     be up to ±20%.  If the response for any analyte varies from the predicted response by more


CD-ROM                               8015B-10                             Revision 2
                                                                         December 1996

-------
     than ±15% (±20% for azeotropic distillation), corrective action must be taken to restore the
     system or a new calibration curve must be prepared for that compound.

          7.5.3  All target analytes and surrogates or n-alkanes in the calibration verification
     analyses must fall within previously established retention time windows. If the retention time
     of any analyte does not fall within the ± 3o window, corrective action must be taken to restore
     the system or a new calibration curve must be prepared for that compound.

          7.5.4  Solvent blanks and any method blanks should be run with calibration verification
     analyses to confirm that laboratory contamination does not cause false positives.

     7.6  Gas chromatographic analysis

          7.6.1  Samples are analyzed in a set referred to  as an analysis  sequence.  The
     sequence  begins with calibration verification followed by sample extract analyses. Additional
     analyses of the verification standard(s) throughout a 12-hour shift are strongly recommended,
     especially for samples that contain visible concentrations of oily material.  A verification
     standard is also necessary at the end of a set. The sequence ends when the set of samples
     has been injected or when retention time and/or % difference QC criteria are exceeded.

          If the criteria are exceeded, inspect the gas chromatographic system to determine the
     cause and perform whatever maintenance is necessary before recalibrating and proceeding
     with sample analysis.  All sample analyses performed using external standard calibration must
     be bracketed with acceptable data quality analyses (e.g., calibration and retention time criteria).
     Therefore, all samples must be reanalyzed that fall within the standard that exceeded criteria
     and the last standard that was acceptable.

          7.6.2  Samples are analyzed with the same instrument configuration as is used during
     calibration. When using Method 5030 for sample introduction,  analysts are cautioned that
     opening a  sample vial or drawing an aliquot from a sealed vial (thus creating headspace) will
     compromise  samples analyzed for volatiles.  Therefore, it is recommended that analysts
     prepare two samples for purge-and-trap analysis.  The second sample can be stored for 24
     hours to ensure that an uncompromised sample is available for analysis or dilution, if the
     analysis of the first sample is unsuccessful  or if results exceed  the calibration range of the
     instrument. Distillates from Method 5031 may be split into two portions and held at 4°C prior
     to analysis. It is recommended that the distillate be analyzed within 24 hours of distillation.
     Distillates must be analyzed within 7 days of distillation.

          7.6.3  Sample concentrations are calculated by comparing sample response data with
     the initial calibration of the system (Sec. 7.3).  Therefore, if sample response exceeds the limits
     of the initial calibration range, a dilution of the sample must be analyzed. For volatile organic
     aqueous samples, the dilution must be performed on a second aliquot of the sample which has
     been properly sealed and stored prior to use and reanalysis. Extracts should be diluted so that
     all peaks are on scale, as overlapping peaks are not always evident when peaks are off scale.
     Computer reproduction of chromatograms, manipulated to ensure all peaks are on scale over
     a 100-fold range, are acceptable as long as calibration limits are not exceeded.  Peak height
     measurements are recommended over peak area integration when overlapping peaks cause
     errors in area integration.

          7.6.4  Tentative identification of a single component analyte occurs when a peak from
     a sample extract falls within the daily retention time window. Confirmation is required on a
     second  column or by GC/MS. Since the flame ionization detector is  non-specific, it is highly


CD-ROM                               8015B -11                              Revision 2
                                                                          December 1996

-------
     recommended that GC/MS confirmation be performed on single component analytes unless
     historical data are available to support the identification(s).

          7.6.5   Second column  confirmation  is  generally  not necessary for petroleum
     hydrocarbon analysis. However, if analytical interferences are indicated, analysis using the
     second GC column is required. Also, the analyst must ensure that the sample hydrocarbons
     fall within the retention time range established during the initial calibration.

          NOTE:    Identification of fuels,  especially gasoline, is complicated  by their inherent
                    volatility. The early eluting compounds in fuels are obviously the most volatile
                    and the most likely to have weathered unless sampled immediately following
                    a spill. The most highly volatile fraction of gasoline constitutes 50% of the
                    total peak area of a gasoline chromatogram.  This fraction is least likely to be
                    present in an environmental sample or present in only very low concentration
                    in relation  to the remainder of a gasoline chromatogram.

          7.6.6   The performance of the entire analytical system should be checked every 12
     hours, using data gathered from analyses of  blanks,  standards, and  replicate samples.
     Significant peak tailing must be corrected. Tailing problems are generally traceable to active
     sites on the column, cold spots  in a GC, the detector operation, or leaks in the system.  See
     Sec. 7.9 for GC/FID system  maintenance. Follow manufacturer's instructions for maintenance
     of the introduction device.

     7.7  Calculations

          7.7.1   The  concentration of each  analyte in the sample  may  be determined by
     calculating the amount of standard purged or injected,  from the peak response, using the
     calibration curve or the mean CF or RF from the initial curve.

          7.7.2   While both diesel fuel and gasoline contain a large number of compounds that
     will produce well resolved peaks in a GC/FID chromatogram, both fuels contain many other
     components that  are not chromatographically resolved.  This unresolved  complex mixture
     results in the "hump" in the chromatogram that  is characteristic of these fuels.  In addition,
     although the resolved peaks are important for the identification of the specific fuel type, the
     area of the unresolved complex mixture contributes a significant portion of the area of the total
     response.

                 7.7.2.1  For the analysis of DROs, sum the area of all peaks eluting between C10
          and C^. This area is generated by projecting a horizontal baseline between the retention
          times of C10  and C28.

                 7.7.2.2  Because the chromatographic conditions employed for DRO analysis
          can result in significant column bleed and a resulting rise in the baseline, it is appropriate
          to perform a subtraction of the column bleed from the area of the DRO chromatogram.
          In order to accomplish this subtraction, a methylene chloride blank should be analyzed
          during each 12-hour analytical shift during which samples are analyzed for DROs. The
          area of this chromatogram is measured in the same fashion as is used for samples (see
          Sec. 7.7.2.1), by projecting a horizontal baseline  across the retention time range for
          DROs.  This area is  then subtracted from the area measured for the sample and the
          difference in areas  is used to calculate the DRO concentration,  using the equations in
          Method 8000.
CD-ROM                               8015B-12                             Revision 2
                                                                         December 1996

-------
                 7.7.2.3  For the analysis of GROs, sum the area of all peaks eluting between 2-
          methylpentane and 1,2,4-trimethyl benzene. This area is used to calculate the GRO
          concentration, using the equations in Method 8000.  Column bleed subtraction is not
          generally required for GRO analysis.

          7.7.3   Refer to Method 8000, Sec. 7.0 for calculation formulae.  The formulae cover
     external and internal standard calibration, aqueous and non-aqueous samples and linear and
     non-linear calibration curves.

     7.8  Screening

          7.8.1   Method 8015 with single-point calibration can also be used for GC/FID screening
     in order to reduce instrument down-time when highly contaminated samples are analyzed using
     GC/MS (e.g., Methods 8260 and 8270).

          7.8.2   The same configuration  of introduction device interfaced to the GC/MS may be
     utilized for the GC/FID or alternative configurations are acceptable.

          7.8.3   Establish that the system response and chromatographic retention times are
     stable. Analyze the high-point GC/MS calibration standard.

          7.8.4   Analyze samples  or sample extracts.  Compare peak heights in the sample
     chromatograms with the high-point standard to establish that no compound with the same
     retention time as a target analyte exceeds the calibration range.  However, the FID is much
     less sensitive  to halogenated compounds than the GC/MS system, therefore, the  above
     comparison is not an absolute certainty.

          7.8.5   It is recommended that the high-point standard should be run at least  every 12
     hours to  confirm the stability  of instrument response and chromatographic retention times.
     However, there is no QC requirement for screening.

     7.9  Instrument Maintenance

          7.9.1   Injection of sample extracts from waste sites often leaves a high boiling residue
     in:  the  injection port area,  splitters when  used,  and  the  injection  port end of  the
     chromatographic column. This residue effects chromatography in many ways (i.e., peak tailing,
     retention time shifts, analyte degradation, etc.) and, therefore, instrument maintenance is very
     important. Residue buildup in a splitter may limit flow through one leg and therefore change
     the split ratios.  If this occurs during an analytical run, the quantitative data may be incorrect.
     Proper cleanup techniques will minimize the problem and instrument QC will indicate when
     instrument maintenance is required.

          7.9.2   Suggested chromatograph maintenance

          Corrective measures may require any one or more of the following remedial actions. Also
     see Sec.  7.0 in Method 8000 for additional guidance on corrective action for capillary columns
     and the injection port.

                 7.9.2.1  Splitter connections - For dual columns which are connected using a
          press-fit  Y-shaped glass splitter or a Y-shaped fused-silica connector,  clean and
          deactivate the splitter or replace with a cleaned and deactivated splitter.  Break off the
          first few inches (up to one foot) of the injection port side of the column.  Remove the


CD-ROM                               8015B -13                             Revision 2
                                                                         December 1996

-------
          columns and solvent backflush according to the manufacturer's instructions.  If these
          procedures fail to eliminate the degradation problem, it may be necessary to deactivate
          the metal injector body and/or replace the columns.

                 7.9.2.2  Column rinsing - The column should be rinsed with several column
          volumes of an appropriate solvent.  Both polar and nonpolar solvents are recommended.
          Depending on the nature of the sample residues expected, the first rinse might be water,
          followed by methanol and acetone; methylene chloride is a satisfactory final rinse and in
          some cases may be the only solvent required. The column should  then be filled with
          methylene chloride and allowed to remain flooded overnight to allow materials within the
          stationary phase to migrate into the solvent.  The column is then flushed with fresh
          methylene chloride, drained, and dried at room temperature with a stream of ultrapure
          nitrogen passing through the column.


8.0  QUALITY CONTROL

     8.1  Refer to Chapter One and Method 8000 for specific quality control (QC) procedures.
Quality control procedures to ensure the proper operation of the various sample preparation and/or
sample introduction techniques can be found in Methods 3500 and 5000.  Each laboratory should
maintain a formal  quality assurance program.  The laboratory should also maintain records to
document the quality of the data generated.

     8.2  Quality control procedures necessary to evaluate the GC system operation are found in
Method 8000, Sec. 7.0 and include evaluation of retention time windows, calibration verification and
chromatographic analysis of samples.

     8.3  Initial Demonstration of Proficiency - Each laboratory must demonstrate initial proficiency
with each sample preparation and determinative method combination it utilizes, by generating data
of acceptable accuracy and precision for target analytes in a clean matrix. The laboratory must also
repeat  the  following operations whenever  new  staff  are trained  or  significant changes in
instrumentation are made. See  Method 8000,  Sec. 8.0 for information on how to accomplish this
demonstration.

     8.4  Sample Quality Control for Preparation and Analysis - The laboratory must also have
procedures for documenting the  effect of the matrix on method performance (precision, accuracy,
and detection limit). At a minimum, this includes the analysis of QC samples including a method
blank, a matrix spike, a duplicate, and a laboratory control sample (LCS) in each analytical batch and
the addition of surrogates to each field sample and QC sample.

          8.4.1   Documenting the effect of the matrix should include the analysis of at least one
     matrix spike and one duplicate unspiked sample or one matrix spike/matrix spike duplicate pair.
     The decision on whether to prepare and analyze duplicate samples or a matrix spike/matrix
     spike duplicate must be based on a knowledge of the samples in the sample batch. If samples
     are expected to contain  target analytes, then laboratories may use one matrix spike and a
     duplicate analysis of an unspiked field sample.  If samples are not expected to contain target
     analytes, laboratories should  use a matrix spike and matrix spike duplicate pair.

          8.4.2   A Laboratory Control Sample (LCS) should be included with each analytical batch.
     The LCS consists of an aliquot of a clean  (control) matrix similar to the sample matrix and of
     the same weight or volume.  The  LCS is spiked with the same analytes at the same
     concentrations as the matrix spike.  When the  results of the matrix spike analysis indicate a


CD-ROM                               8015B-14                             Revision 2
                                                                         December 1996

-------
     potential problem due to the sample matrix itself, the LCS results are used to verify that the
     laboratory can perform the analysis in a clean matrix.

          8.4.3  See Method 8000, Sec. 8.0 for the details on carrying out sample quality control
     procedures for preparation and analysis.

     8.5  Surrogate recoveries - The laboratory must evaluate surrogate recovery data from
individual samples versus the surrogate control limits developed by the laboratory. See Method
8000, Sec. 8.0 for information on  evaluating  surrogate data and developing and updating surrogate
limits.

     8.6  It is recommended that the laboratory adopt additional quality assurance practices for use
with this method. The specific  practices that are most productive depend upon the needs of the
laboratory and  the nature of the samples. Whenever possible, the laboratory should analyze
standard reference materials and participate in relevant performance evaluation studies.


9.0  METHOD  PERFORMANCE

     9.1  Specific method performance information for non-purgeable volatiles prepared using the
azeotropic microdistillation technique from Method 5031 is included in Tables 1, 3 and 4 for aqueous
matrices and in  Tables 2 and 5 for solid matrices.

     9.2  Specific method performance information is provided for diesel fuel spiked into soil in
Tables 6 and 7.
10.0 REFERENCES

1.    Bellar, T.A., and J.J. Lichtenberg. "Determining Volatile Organics at Microgram-per-Liter Levels
     by Gas Chromatography", J. Amer. Water Works Assoc., 66(12V pp. 739-744 (1974).

2.    Bellar, T.A., and J.J. Lichtenberg. "Semi-Automated Headspace Analysis of Drinking Waters
     and  Industrial Waters for  Purgeable Volatile  Organic  Compounds", in Van  Hall, ed.,
     Measurement of Organic Pollutants in Water and Wastewater, ASTM STP 686, pp. 108-129,
     1979.

3.    Development and Application of Test Procedures for Specific Organic Toxic Substances in
     Wastewaters:  Category 11 -  Purgeables  and  Category 12 - Acrolein,  Acrylonitrile, and
     Dichlorodifluoromethane, Report for EPA Contract 68-03-2635.

4.    Bruce, M.L., R.P. Lee, and M.W. Stevens. "Concentration of Water Soluble Volatile Organic
     Compounds from Aqueous Samples by Azeotropic Microdistillation", Environ. Sci. Technol.
     1992, 26., 160-163.

5.    Tsang, S.F., N. Chau, P.J. Marsden, and K.R. Carter. "Evaluation of the EnSys PETRO RISc.
     kitforTPH", Report for Ensys, Inc., Research Triangle Park, NC, 27709,1992.

6.    "Intertaboratory Study of Three Methods for Analyzing Petroleum Hydrocarbons in Soils," API
     Publication Number 4599, American Petroleum Institute, March 1994.
CD-ROM                              8015B -15                             Revision 2
                                                                        December 1996

-------
                                  TABLE 1

     METHOD DETECTION LIMITS FOR NON-PURGEABLE VOLATILE COMPOUNDS
    IN AQUEOUS MATRICES BY AZEOTROPIC MICRODISTILLATION (METHOD 5031)
Analyte
Acetone"
Acetonitrile
Acrolein
Acrylonitrile
1-Butanol
t-Butyl alcohol
1 ,4-Dioxane
Ethanol
Ethyl acetate
Ethylene oxide
Isobutyl alcohol
Isopropyl alcohol
Methanol
Methyl ethyl ketone
Methyl isobutyl ketone
2-Pentanone
1-Propanol
Propionitrile
Pyridine

Reagent Water
48
15
13
8
14
8
12
18
9
8
11
18
21
4
4
2
—
10
11
MDL (ua/L)a
Ground Water
16
6
15
9
8
7
15
12
8
9
8
17
21
5
2
2
7
6
9

TCLP Leachate
63
14
7
14
7
17
16
13
16
10
4
7
22
9
8
7
—
13
21
 "    Produced by analysis of 7 aliquots of water spiked at 25 UQ/L. using internal standard
      calibration.
 b    Problematic due to transient laboratory contamination.
CD-ROM                          8015B-16                          Revision 2
                                                               December 1996

-------
                                   TABLE 2

      METHOD DETECTION LIMITS FOR NON-PURGEABLE VOLATILE COMPOUNDS
      IN SOLID MATRICES BY AZEOTROPIC MICRODISTILLATION (METHOD 5031)
                                                      MDL fma/kcri
      Analyte                              Incinerator Ash            Kaolin
Acrylonitrile
1-Butanol
t-Butyl alcohol
1,4-Dioxane
Ethanol
Ethyl acetate
Isopropyl alcohol
Methanol
Methyl ethyl ketone
Methyl isobutyl ketone
2-Pentanone
Pyridine
0.42
0.23
0.34
0.31
0.47
0.18
0.40
0.46
0.27
0.12
0.16
0.20
0.09
0.09
0.13
0.16
0.19
0.07
0.19
0.31
0.12
0.05
0.07
0.08
The MDLs calculated for this table were produced by the analysis of 7 replicates spiked at 0.50
mg/kg, using internal standard calibration.
CD-ROM                           8015B -17                          Revision 2
                                                                 December 1996

-------
                               TABLE 3

    METHOD PERFORMANCE DATA FOR NON-PURGEABLE VOLATILES IN GROUND
           WATER BY AZEOTROPIC MICRODISTILLATION (METHOD 5031)


Compound
Acetone*
Acetonitrile
Acrolein
Acrylonitrile
1-Butanol
t-Butyl alcohol
1 ,4-Dioxane
Ethanol
Ethyl Acetate
Ethylene oxide
Isobutyl alcohol
Isopropyl alcohol
Methanol
Methyl ethyl ketone
Methyl isobutyl ketone
2-Pentanone
1-Propanol
Propionitrile
Pyridine
Low Gone.8
Average"
%Rec %RSD
126 17
147 5
146 13
179 7
127 8
122 7
124 16
152 10
142 7
114 10
122 8
167 13
166 14
105 6
66 4
94 3
N/A
135 5
92 12
Medium Cone."
Average"
%Rec %RSD
N/A
105
120
143
86
N/A
96
N/A
135
N/A
87
.N/A
94
N/A
N/A
N/A
91
102
N/A

8
27
28
8
—
10
—
33
—
13
—
9
—
—
—
7
14
~~
High Conc.c
Average"
%Rec
N/A
92
80
94
90
N/A
99
N/A
92
N/A
89
N/A
95
N/A
N/A
N/A
91
90
N/A
%RSD

9
20
21
9
—
8
~
25
—
13
~
7
—
—
—
7
14
«.
8 25 ug/L spikes, using internal calibration.
b 100 ug/L spikes,
c 750 pg/L spikes,
using internal calibration.
using internal calibration.








6 Average of 7 replicates
e Problematic due
to transient laboratory contamination.
 N/A  Data not available
CD-ROM
8015B-18
    Revision 2
December 1996

-------
                                TABLE 4

     METHOD PERFORMANCE DATA FOR NON-PURGEABLE VOLATILES IN TCLP
         LEACHATE BY AZEOTROPIC MICRODISTILLATION (METHOD 5031)


Compound
Acetone8
Acetonitrile
Acrolein
Acrylonitrile
1-Butanol
t-Butyl alcohol
1 ,4-Dioxane
Ethanol
Ethyl Acetate
Ethylene oxide
Isobutyl alcohol
Isopropyl alcohol
Methanol
Methyl ethyl ketone
Methyl isobutyl ketone
2-Pentanone
1-Propanol
Propionitrile
Pyridine
Low Cone.8
Average1'
%Rec %RSD
99 91
107 17
88 10
133 13
119 7
70 31
103 20
122 13
164 12
111 12
115 4
114 8
107 10
87 13
78 13
101 8
N/A
100 16
46 59
Medium Cone."
Average"
%Rec
N/A
111
109
123
89
N/A
103
N/A
119
N/A
86
N/A
102
N/A
N/A
N/A
98
100
N/A
%RSD

10
29
29
12
~
16
—
29
~
13
~
6
~
~
—
10
11
•"•
High Conc.c
Average"
%Rec
N/A
95
87
103
86
N/A
102
N/A
107
N/A
82
N/A
N/A
N/A
N/A
N/A
89
90
N/A
%RSD

11
41
38
8
—
7
—
41
—
13
—
—
—
—
—
7
17
~—
a 25 ug/L spikes, using internal calibration.
b 100 ug/L spikes,
c 750 ug/L spikes,
using internal calibration.
using internal calibration.








d Average of 7 replicates
8 Problematic due
to transient laboratory contamination.
N/A  Data not available
CD-ROM
8015B-19
    Revision 2
December 1996

-------
                               TABLE 5

    METHOD PERFORMANCE DATA FOR NON-PURGEABLE VOLATILE COMPOUNDS
      IN SOLID MATRICES BY AZEOTROPIC MICRODISTILLATION (METHOD 5031)
Incinerator Ash
Low Cone.8 High Cone."
Average0 Average0
%Rec %RSD %Rec
Acrylonitrile
1-Butanol
t-Butyl alcohol
1 ,4-Dioxane
Ethanol
Ethyl acetate
Isopropyl alcohol
Methanol
Methyl ethyl ketone
Methyl isobutyl ketone
2-Pentanone
Pyridine
8 0.5 mg/kg spikes
b OC rvm/lx^i t?v\ilsr\e*
50
105
101
106
117
62
119
55
81
68
79
52
, using
i i*»!rt/*» i
53
14
21
19
25
19
21
53
21
11
13
24
internal
*"» + *"M»*"* « 1
10
61
60
48
52
39
61
33
40
57
54
44
%RSD
31
12
13
18
20
12
15
28
12
14
10
20
Kaolin
Low Cone.8
Average0
%Rec
102
108
97
105
108
90
108
117
91
71
91
50
High Cone."
Average0
%RSD %Rec
6
5
9
10
11
5
11
17
8
5
5
10
12
58
59
48
48
41
58
37
42
55
54
49
%RSD
52
25
23
25
24
25
24
22
20
23
19
31
calibration.
^^ \i\^**%btf\n
     Average of 7 replicates
CD-ROM
8015B-20
    Revision 2
December 1996

-------
                                     TABLE 6

          RESULTS FROM ANALYSIS8 OF LOW AROMATIC DIESEL" BY GC/FID
                                  (5 replicates/test)
Spike Concentration
12.5 ppm
75 ppm
105 ppm
150 ppm
1000 ppm
Analysis Results
ND
54 ± 7 ppm
90 ± 15 ppm
125 ± 12 ppm
960 ± 1 05 ppm
             Samples were prepared using 2 g aliquots of sandy loam soil spiked with known
             amounts of low aromatic diesel.  Extractions were accomplished using methylene
             chloride as a solvent (Method 3550, high concentration option).

             Low aromatic diese! is sold in California (Section 2256, CCR).  For this study it was
             purchased at a gas station in San Diego, California.
                                      TABLE 7

          RESULTS FROM ANALYSIS8 OF LOW AROMATIC DIESEL" BY GC/FID
                                  (5 replicates/test)
Spike Concentration
25 ppm
75 ppm
125 ppm
150 ppm
Analysis Results
51. 2 ±6.4 ppm
75.9 ± 7.8 ppm
98.9 ± 5.2 ppm
162± 10.4 ppm
             Samples were prepared using 10 g aliquots of sandy loam soil spiked with known
             amounts of regular #2 diesel purchased  at a gas station in Northern Virginia.
             Extractions were accomplished using methylene chloride as a solvent (Method 3550).
CD-ROM
8015B-21
     Revision 2
December 1996

-------
                                 CHROMATOGRAM OF A 300 PPM GASOLINE STANDARD
    80000
    60000
    40000
    20000
                                     SF;
                                                                                     ."JUW
                                                                                        £.
                1 • i ' ' ' ' i ' '
                 2.00  3.00
                             N  5
                            jU
 •n"1
5.00
\ i | i i i i'| i i i r^-rn
 6.QO   7.00   8.00
                      It!
                      fie. . A
                         e
                         m
9.00  10.00 11.00 12.00  13.00  14.00  15.00  16.00  17.00 18.00 19.00  20.00  21.00
Time  0.00
1 ' I '
 1.00
1 > i ' '
 4.00
CD-ROM
                        8015B-22
                                                                                                             Revision 2
                                                                                                        December 1996

-------
                                 FIGURE 2

                CHROMATOGRAM OF A 30 PPM DIESEL STANDARD
   abundance
      35000-
      30000-
      25000-
      20000-
      15000-
      10000-
           '.
       5000-
  TIC: 523002.D
          0-
                                            o-Terphenyl
                                            (surrogate)
   rime-->0.00     5.00    10.00    15.00    20.00    25.00    30.00
CD-ROM
8015B-23
    Revision 2
December 1996

-------
                              FIGURE 3

          CHROMATOGRAM OF A 30 PPM DIESEL STANDARD WITH THE
                BASELINE PROJECTED BETWEEN C10 AND C18
                        10.00
        20.00
CD-ROM
8015B-24
    Revision 2
December 1996

-------
                                   FIGURE 4

      CHROMATOGRAM OF SEVERAL NONPURGEABLE VOLATILE COMPOUNDS IN
   SPIKED REAGENT WATER USING AZEOTROPIC MICRODISTILLATION (METHOD 5031)
                  OUBdOjdOS|OJOn|JBX6l| 'S'l
                     louejnq-1. i
                  louedojd-(.-
                                           8U6XO|p-t' I
                                                             J
                     euojaoe
                                                                — 10
        Mix 1:

  Conditions:
Analytes distilled at 0.25mg/L, Internal Stds. at 2.5 mg/L

J&W DB-Wax column with 0.53 ID
Temperature program: 30°C for 2 min.
                     3°C/min. to 100°C and held for 0 min.
                   25°C/min. to 200°C and held for 4 min.
CD-ROM
               8015B-25
    Revision 2
December 1996

-------
                                  FIGURE 5

      CHROMATOGRAM OF SEVERAL NONPURGEABLE VOLATILE COMPOUNDS IN
   SPIKED REAGENT WATER USING AZEOTROPIC MICRODISTILLATION (METHOD 5031)
           Mix 2:       Analytes distilled at 0.25mg/L, Internal Stds. at 2.5 mg/L

     Conditions:        J&W DB-Wax column with 0.53 ID
                      Temperature program: 30°C for 2 min.
                                         3°C/min. to 100°C and held for 0 min.
                                        25°C/min. to 200°C and held for 4 min.
CD-ROM
8015B-26
                                                                   Revision 2
                                                               December 1996

-------
                                            METHOD 8015B
                          NONHALOGENATED ORGANICS USING GC/FID
1
7.2
f
Set chromatographic conditions.
           7.3.2 and 7.3.3 SBO Method
              8000 for guidance on
           external standard calibration.
                                          7.3
                                         Initial
                                      Calibration;
                                       Internal or
                                       External?
                                                   Analysis
                                                 lor individual
                                                  analytes or
                                                    fuels?
    7.3.3.1 Prepare 5
initial calibration standards
    for each fuel type.
             7.3.3.2 Introduce each
             standard into GC using
            technique to be used for
            actual samples. Calculate
                     CF.
 7.3 1  See Method 8000
 for guidance on internal
   standard calibration.
    7.3.2.1 Prepare 5
initial calibration standards
 containing each analyta
       of interest
                                                                 7.3.2.2 Introduce each
                                                                 standard into GC using
                                                                technique to be used for
                                                                actual samples. Calculate
                                                                         CF.
                                                            < 20% over
                                                            the working
                                                               range?
                                        What is
                                      the percent
                                      relative std.
                                       deviation
                                         of the
                                          CF?
                                                                               7.3.4.1 Use the
                                                                             average calibration
                                                                              (actor from initial
                                                                                 calibration.
                                          > 20% over
                                          the working
                                            range?
                                              7.3.4.2 See Method
                                                 8000 for other
                                               calibration options.
CD-ROM
                                   8015B-27
                    Revision 2
              December 1996

-------
                                                 METHOD 8015B
                                                     (continued)
7.4 Establish RT windows for
individual compounds or
establish RT ranges for
GROs and DROs.
>
t
7.6.1 Perform
chromatographic
analysis of samples.
       Recalibrate instruments,
          see Method 8000.
                                     Check system and
                                   reanalyze.  Saa Method
                                    8000 for additional
                                        guidance.
                   7.6.3
                Does sample
               response exceed
                limits of initial
                 calibration?
                                                                                             7.6.3 Dilute sample
                                                                                               and reanalyze.
                                                                               Individual
                                                                     Analysis  \ Analytes
                                                                   of individual
                                                                    analytes or
                                                                     fuels?
                                          7.6.4 Confirm ID of
                                         individual compounds
                                          on a 2nd column or
                                               GC/MS.
  7.6.5 If analytical
  interferences are
indicated, analyze on a
    2nd column.
CD-ROM
8015B-28
                                                                                                        Revision 2
                                                                                                 December 1996

-------
00
o

-------
                                 METHOD 8020

                         AROMATIC VOLATILE ORGANICS


1.0  SCOPE AND APPLICATION

     1.1  Method 8020  1s  used  to  determine  the  concentration  of various
aromatic volatile organic compounds.  Table 1 Indicates compounds which may be
determined by this  method  and  lists  the  method  detection  limit for each
compound 1n reagent water.    Table  2  lists the practical quantltation limit
(PQL) for other matrices.


2.0  SUMMARY OF METHOD

     2.1  Method 8020 provides chromatographlc conditions for the detection of
aromatic volatile compounds.  Samples  can  be analyzed using direct Injection
or purge-and-trap (Method  5030).    Ground  water  samples must be determined
using Method 5030.  A temperature program   1s used 1n the gas chromatograph to
separate the organic compounds.   Detection  1s achieved by a photo-1on1zation
detector  (PID).

     2.2  If Interferences are   encountered,  the  method provides an optional
gas  chromatographlc column that  may be  helpful 1n resolving the analytes  from
the  interferences and for analyte confirmation.


3.0  INTERFERENCES

     3.1  Refer  to Method 5030 and  8000.

     3.2  Samples  can   be   contaminated   by  diffusion  of  volatile organics
(particularly  chlorofluorocarbons and  methylene   chloride) through  the  sample
container septum during  shipment and   storage.   A field sample blank prepared
from reagent water and   carried  through   sampling and subsequent storage and
handling  can serve as a  check on such  contamination.


4.0  APPARATUS AND MATERIALS

     4.1  Gas  chromatograph:

          4.1.1  Gas  Chromatograph:    Analytical  system  complete  with gas
     chromatograph suitable  for  on-column   injections or purge-and-trap  sample
     introduction and all   required  accessories,  Including detectors,  column
     supplies,  recorder,  gases,  and  syringes.     A  data  system  for measuring
     peak heights and/or peak areas is  recommended.
                                   8020 - 1
                                                          Revision       0
                                                          Date   September  1986

-------
TABLE 1.  CHROMATOGRAPHIC CONDITIONS AND METHOD DETECTION LIMITS FOR AROMATIC
          VOLATILE ORGANICS



Compound
Benzene
Chlorobenzene
1 ,4-Dichlorobenzene
1 , 3-Dichlorobenzene
1,2-Dichlorobenzene
Ethyl Benzene
Toluene
Xylenes
Retention
(m1n)

Col. 1
3.33
9.17
16.8
18.2
25.9
8.25
5.75

time


Col. 2
2.75
8.02
16.2
15.0
19.4
6.25
4.25

Method
detection
limit3
(ug/L)
0.2
0.2
0.3
0.4
0.4
0.2
0.2

     a  Using  purge-and-trap  method  (Method  5030).
 TABLE 2.  DETERMINATION OF PRACTICAL QUANTITATION LIMITS (PQL) FOR VARIOUS
           MATRICES3


     Matrix                                                    Factorb


 Ground water                                                     10
 Low-level soil                                                   10
 Water miscible liquid waste                                     500
 High-level soil and sludge                                     1250
 Non-water miscible waste                                       1250


      aSample  PQLs are highly  matrix-dependent.    The  PQLs listed herein are
      provided for guidance and may not always be achievable.

      bPQL = [Method detection limit (Table 1)] X [Factor (Table 2)].  For non-
      aqueous  samples, the factor 1s on a wet-weight basis.
                                   8020 - 2
                                                          Revision      0
                                                          Date  September 1986

-------
          4.1.2  Columns:

               4.1.2.1  Column 1:   6-ft  x  0.082-1n I.D. 1304 stainless steel
          or glass column  packed   with  5%  SP-1200  and  1.75% Bentone-34 on
          100/120 mesh Supelcort or equivalent.

               4.1.2.2  Column 2:   8-ft x O.l-1n I.D. stainless steel  or glass
          column packed with 5% 1,2,3-Tr1s(2-cyanoethoxy)propane on 60/80 mesh
          Chromosorb W-AW or equivalent.

          4.1.3  Detector:  Photoion1zat1on  (PID)   (h-Nu  Systems, Inc. Model
     PI-51-02 or equivalent).

     4.2  Sample  introduction  apparatus;    Refer  to  Method  5030  for the
appropriate equipment for sample introduction purposes.

     4.3  Syringes;  A 5-mL  Luerlok  glass  hypodermic  and a 5-mL, gas-tight
with shutoff valve.

     4.4  Volumetric flask:  10-,  50-, 100-, 500-,  and 1,000-mL with a ground-
glass stopper.

     4.5  Microsyringe;   10- and 25-uL  with  a 0.006-1n I.D. needle (Hamilton
702N or equivalent) and  a 100-uL.
5.0  REAGENTS

     5.1  Reagent water;  Reagent  water  1s  defined  as
interferent  is  not  observed  at  the  method  detection
parameters of  interest.
a water in which an
limit  (MOL) of the
      5.2  Stock  standards:   Stock  solutions may be prepared  from pure  standard
materials or .purchased  as   certified   solutions.     Prepare  stock  standards  in
methanol using assayed  liquids.  Because   of   the toxlclty of  benzene  and  1,4-
dichlorobenzene,  primary dilutions of  these   materials  should  be prepared  in  a
hood.

           5.2.1   Place  about 9.8 ml of methanol  1n a 10-mL tared ground-glass-
      stoppered volumetric  flask.   Allow  the   flask  to  stand,  unstoppered,  for
      about  10 min or until  all  alcohol-wetted surfaces have dried.  Weigh  the
      flask  to the nearest  0.1 mg.

           5.2.2   Using  a  100-uL syringe,  Immediately add two  or more  drops  of
      assayed  reference  material to the  flask;  then rewelgh.   The liquid must
      fall  directly into the alcohol without  contacting  the neck of the flask.

           5,2.3   Reweigh,  dilute  to volume,  stopper, and then  mix  by inverting
      the  flask  several  times.  Calculate the concentration in micrograms  per
      microliter  (ug/uL) from the  net  gain  in weight.   When  compound purity  is
      assayed  to  be 96%  or  greater, the  weight may  be  used  without  correction
                                   8020 - 3
                                                          Revision      0
                                                          Date  September  1986

-------
     to  calculate   the   concentration   of   the   stock   standard.   Commercially
     prepared stock standards  may   be   used  at   any concentration if  they  are
     certified by  the manufacturer  or  by an independent source.

          5.2.4  Transfer the   stock  standard   solution  into  a  Teflon-sealed
     screw-cap bottle.   Store, with minimal headspace,  at 4*C and protect  from
     light.

          5.2.5  All standards must be replaced   after   6 months, or sooner if
     comparison with check standards indicates  a problem.

     5.3  Secondary dilution standards:   Using  stock  standard solutions,  pre-
pare in methanol secondarydilution  standards,  as   needed, that contain the
compounds of  interest,  either  singly  or  mixed  together.    The secondary
dilution standards should be prepared  at  concentrations such that the aqueous
calibration standards prepared in  Paragraph 5.4  will  bracket the  working range
of the analytical  system.  Secondary  dilution  standards should be stored  with
minimal  headspace for volatiles and should  be  checked  frequently for signs of
degradation or evaporation,  especially  just  prior  to preparing calibration
standards from them.

     5.4  Calibration standards;  Calibration  standards  at a minimum of five
concentration levels are prepared in reagent water from the secondary dilution
of the stock  standards.    One  of  the  concentration  levels  should be at a
concentration near, but  above,  the  method  detection  limit.  The remaining
concentration levels should correspond to the expected range of concentrations
found in real samples or   should  define  the  working  range of the GC.  Each
standard should contain each  analyte for  detection by this method  (e.g.,  some
or all of "the compounds  listed  in  Table  1  may  be included).   In order to
prepare accurate aqueous standard solutions, the following precautions must be
observed.

          5.4.1  Do not  inject more than 20 uL  of alcoholic  standards Into 100
     ml of  reagent  water.

          5.4.2  Use   a  25-uL   Hamilton   702N  microsyringe   or  equivalent
      (variations  in  needle   geometry  will  adversely   affect   the ability to
     deliver reproducible  volumes of methanolic  standards  into water).

           5.4.3  Rapidly  inject   the  alcoholic   standard   into  the  filled
      volumetric flask.   Remove  the  needle  as fast  as possible  after injection.

           5.4.4  Mix aqueous   standards   by  inverting  the   flask  three  times
      only.

           5.4.5  Fill  the  sample  syringe   from  the standard  solution  contained
      in the expanded area  of  the   flask   (do not use any solution contained  in
      the neck of the flask).
                                   8020 - 4
                                                          Revision      0
                                                          Date  September  1986

-------
          5.4.6  Never use pipets  to  dilute  or  transfer samples or aqueous
     standards.

          5.4.7  Aqueous standards  are  not  stable  and  should be discarded
     after 1 hr,  unless properly sealed and stored.  The aqueous standards  can
     be stored up to 24 hr, if held in sealed vials with zero headspace.

     5.5  Internal standards (If Internal  standard  calibration 1s used):   To
use this approach, the analyst must select one or more internal  standards that
are similar in analytical behavior to  the compounds of Interest.  The analyst
must further demonstrate that the measurement  of the Internal standard is not
affected by method or matrix interferences.   Because of these limitations, no
Internal standard can be suggested  that  is  applicable  to all samples.  The
compound,  alpha,alpha,alpha-trifluorotoluene   recommended   for   use  as  a
surrogate spiking compound (Paragraph  5.6)  has  been used successfully as an
internal standards.

          5.5.1  Prepare  calibration   standards   at   a   minimum  of  five
     concentration levels  for  each  parameter  of  interest  as described in
     Section 5.4.

          5.5.2  Prepare a spiking  solution  containing  each of the internal
     standards using the procedures described In  Sections 5.2 and 5.3.  It is
     recommended  that  the  secondary  dilution  standard  be  prepared  at a
     concentration of   15  ug/mL  of  each  Internal  standard  compound.  The
     addition  of 10 uL of  this  standard  to  5.0 ml of sample or calibration
     standard  would be equivalent to 30 ug/L.

          5.5.3  Analyze each calibration  standard  according to Section 7.0,
     adding 10  uL  of  internal  standard  spiking  solution  directly to the
     syringe.

     5.6  Surrogate  standards;     The   analyst   should  monitor  both  the
performance of the analytical system  and  the  effectiveness of the method in
dealing with each sample matrix by  spiking each sample, standard, and reagent
water  blank with  surrogate compounds  (e.g, alpha,alpha,alpha-trifluorotoluene)
recommended to encompass  the  range  of  the  temperature program used in this
method.   From  stock  standard   solutions  prepared  as   in  Section 5.2, add a
volume to give 750 ug  of  each surrogate  to 45 ml of reagent water contained in
a  50-mL volumetric flask,  mix,  and dilute  to  volume for  a concentration  of
15 ng/uL.   Add 10 uL of this  surrogate spiking solution  directly into  the  5-mL
syringe with every sample  and   reference   standard  analyzed.   If the  internal
standard  calibration procedure  is  used,  the1  surrogate  compounds may  be added
directly  to the  internal  standard  spiking  solution  (Paragraph 5.5.2).

     5.7  Methanol;   pesticide  quality or   equivalent.   Store  away  from other
solvents.
                                   8020 - 5
                                                          Revision       0
                                                          Date   September  1986

-------
6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1  See the Introductory  material  to  this  chapter, Organic Analytes,
Section 4.1.


7.0  PROCEDURE

     7.1  Volatile compounds are  Introduced   Into the gas chromatograph either
by direct Injection or purge-and-trap  (Method  5030).  Method 5030 may be used
directly  on  ground   water  samples    or  low-level  contaminated   soils  and
sediments.  For  medium-level   soils   or  sediments, methanollc  extraction, as
described 1n Method 5030, may be  necessary prior to purge-and-trap analysis.

     7.2  Gas chromatography conditions (Recommended);

          7.2.1  Column  1:  Set helium gas   flow  at 36  mL/m1n flow  rate.  The
     temperature program  sequences   are  as  follows:     For   lower  boiling
     compounds,  operate  at  50*C Isothermal for  2 m1n; then program  at 6*C/m1n
     to 90*C and hold until  all   compounds   have  eluted.   For  higher boiling
     range  of compounds,  operate  at 50*C  Isothermal  for  2 m1n; then  program at
     3*C/m1n to  110*C and  hold  until   all   compounds   have eluted.  Column  1
     provides  outstanding   separations  for   a   wide  variety of aromatic
     hydrocarbons.   Column  1 should  be  used as the primary analytical column
     because of  Its  unique  ability to resolve para-, meta-, and  ortho-aromatic
     Isomers.

           7.2.2  Column 2:   Set helium gas   flow  at 30  mL/m1n flow  rate.  The
     temperature program sequence 1s  as  follows:    40*C Isothermal  for 2 nrin;
     then 2*C/m1n  to  100'C  and  hold  until   all compounds have eluted.  Column
     2, an  extremely  high-polarity column, has  been  used for a number of years
     to  resolve  aromatic   hydrocarbons   from  alkanes  1n  complex samples.
     However, because resolution   between some of  the aromatlcs   1s not as
     efficient  as  with Column  1,   Column   2   should  be  used as  a confirmatory
     column.

     7.3  Calibration;    Refer   to   Method  8000   for   proper  calibration
 techniques. Use Table 1 and especially  Table  2 for guidance on selecting the
 lowest point on  the calibration curve.

           7.3.1   Calibration  must   take   place    using    the   same  sample
      Introduction  method that   will  be  used  to   analyze  actual samples  (see
      Section 7.4.1).

           7.3.2  The procedure for  Internal  or  external   calibration may  be
      used.     Refer  to  Method  8000  for  a  description   of   each  of  these
      procedures.
                                   8020 - 6
                                                          Revision
                                                          Date  September 1986

-------
    7.4  Gas chromatographic analysis;

         7.4.1  Introduce volatile compounds Into the gas chromatograph using
    either  Method  5030  (purge-and-trap  method)  or  the  direct injection
    method.  If the internal standard calibration technique is used, add
    10 uL of internal standard to the sample prior to purging.

              7.4.1.1  Direct Injection;  In very limited applications (e.g.,
         aqueous process wastes), direct Injection  of the sample Into the GC
         system with a 10 uL syringe may be appropriate.  The detection limit
         is very high  (approximately  10,000  ug/L);  therefore,  it is only
         permitted when concentrations in excess  of 10,000 ug/L are expected
         or for water-soluble compounds that  do  not purge.  The system must
         be calibrated  by  direct  Injection  (bypassing  the purge-and-trap
         device).

         7.4.2  Follow Section 7.6  of  Method  8000  for Instructions on the
    analysis  sequence,  appropriate   dilutions,   establishing dally retention
    time windows,  and identification  criteria.    Include a mid-level standard
    after  each group of  10  samples 1n the analysis sequence.

         7.4.3  Table   1   summarizes    the   estimated   retention   times  and
    detection limits for  a  number of organic compounds  analyzable  using  this
    method.   An example  of  the   separation   achieved  by Column  1  1s  shown  in
    Figure 1.   Figure 2   shows   an   example of  the  separation achieved using
    Column 2.

         7.4.4  Record  the  sample volume purged  or Injected  and  the resulting
    peak sizes  (1n area  units or peak heights).

         7.4.5  Calculation of concentration  1s  covered   ir Section  7.8  of
    Method 8000.

          7.4.6   If analytical  interferences are  suspected,  or for  the  purpose
    of confirmation, analysis  using  the second GC column  1s  recommended.

          7.4.7   If the  response  for  a  peak  is  off-scale,  prepare a  dilution
    of the sample with  reagent water.    The  dilution  must  be performed  on a
     second aliquot of  the sample which  has  been properly  sealed and  stored
    prior  to use.


8.0  QUALITY CONTROL

     8.1  Refer to Chapter  One   for   specific  quality  control procedures and
Method 8000 for gas chromatographic  procedures.   Quality control  to ensure the
proper operation of the purge-and-trap device is  covered in  Method  5030.

     8.2  Mandatory quality control   to   validate  the  GC system operation  is
found  in Method 8000, Section 8.6.
                                  8020 - 7
                                                         Revision      0
                                                         Date  September 1986

-------
                                                                               Column: 6% SP-1200/1.75% Bottom 34

                                                                               Program. 60°C-2 Minum. 6°C/Min. to 00°C

                                                                               D«ttctor: Phototoniiation

                                                                               Sample: 0.40 pg/1 Standard Mixture
     CO
     o
     rv>
     o
     CO
   n>
GO O
fO  3
cr
                                                        a        to       12       M

                                                         RETENTION TIME (MINUTES)
16
18
20
vo
CO
                                Figure 1.  Chromatogram of aromatic volatile organic* (column 1 conditions).

-------
                                 Column: S* 1 JJ-Trii (2-Cvanecthoxy)
                                 Prooww on Chromotorb—W
                                 Program: 40°C-2 MinutM 2»C/M»n. to 100°C
                                 Ovnetor: fttotoionization
                                 Scmpit: 2.0^8/1 Sundard Mixturt
                        I          12         16

                      RETENTION TIME (MINUTES)
24
Figurt 2. Chrprwogrim of iromf tic volatile organic* (column 2 conditioru).
                        8020 -  9
                                                  Revision       0
                                                  Date   September  1986

-------
          8.2.1  The quality control   check  sample  concentrate (Method 8000,
     Section 8.6)  should contain each parameter of Interest at a concentration
     of 10 ug/mL in methanol.

          8.2.2  Table 3 indicates the  calibration and QC acceptance criteria
     for this  method.    Table  4  gives  method  accuracy  and  precision as
     functions of concentration for the analytes of interest.  The contents of
     both Tables should be used to  evaluate a laboratory's ability to perform
     and generate acceptable data by this method.

     8.3  Calculate surrogate standard  recovery  on  all samples, blanks,  and
spikes.   Determine  if  recovery  is  within  limits  (limits  established by
performing QC procedure outlined in Method 8000, Section 8.10).

          8.3.1  If recovery is not within limits, the following is required.

               •  Check to  be  sure  there  are  no  errors  in calculations,
                  surrogate solutions  and  Internal  standards.   Also, check
                  instrument performance.

               •  Recalculate the data and/or reanalyze  the extract if any of
                  the above checks reveal a problem.

               •  Reextract and reanalyze the sample  1f none of the above are
                  a problem or  flag the data as  "estimated concentration."


9.0  METHOD  PERFORMANCE

     9.1  This method  was   tested  by  20  laboratories   using  reagent water,
drinking  water, surface water,  and  three  industrial wastewaters  spiked at six
concentrations over  the  range  2.1-500   ug/L.     Single  operator  precision,
overall  precision,  and method  accuracy  were   found to  be directly  related  to
the  concentration  of  the  parameter  and  essentially Independent  of  the sample
matrix.   Linear equations  to   describe  these   relationships  are presented  in
Table  4.

     9.2  The accuracy  and  precision  obtained  will  be determined by  the sample
matrix,  sample introduction technique,  and by  the calibration  procedure used.


 10.0  REFERENCES


 1.   Bellar,  T.A.,  and J.J.  Lichtenberg,   J.  Amer.  Water Works Assoc.,  66(12),
pp.  739-744, 1974.

 2.   Bellar,  T.A.,  and J.J.   Lichtenberg,   Semi-Automated Headspace Analysis  of
 Drinking  Waters   and  Industrial   Waters   for  Purgeable  Volatile  Organic
Compounds,  in Van  Hall  (ed.),   Measurement  of Organic  Pollutants in Water and
Wastewater,  ASTM  STP 686, pp.  108-129,  1979.
                                   8020 - 10
                                                          Revision      0
                                                          Date  September 1986

-------
3.  Dowty, B.J., S.R. Antolne, and J.L. Laseter, "Quantitative and Qualitative
Analysis of Purgeable Organlcs by High Resolution Gas Chromatography and Flame
lonization Detection," 1n Van Hall,  ed., Measurement of Organic Pollutants 1n
Water and Wastewater.  ASTM STP 686, pp. 24-35, 1979.

4.  Development and Application of Test  Procedures for Specific Organic Toxic
Substances in  Wastewaters.    Category  11  -  Purgeables  and  Category 12 -
Acroleln, Acrylon1tr1le, and D1chlorod1fluoromethane.  Report for EPA Contract
68-03-2635 (1n preparation).

5.  "EPA Method Validation Study 24, Method 602 (Purgeable Aromatlcs)," Report
for EPA Contract 68-03-2856  (in preparation).

6.  U.S.  EPA 40 CFR  Part 136, "Guidelines Establishing Test Procedures for the
Analysis  of Pollutants Under  the Clean Water Act; Final Rule and Interim Final
Rule and  Proposed Rule," October 26, 1984.

7.  Provost, L.P., and R.S.  Elder,   "Interpretation of Percent Recovery Data,"
American  Laboratory,  1_5, pp.  58-63,  1983.
                                   8020 - 11
                                                          Revision
                                                          Date  September 1986

-------
TABLE 3.  CALIBRATION AND QC ACCEPTANCE CRITERIA3


Parameter
Benzene
Chlorobenzene
l,2-D1chlorobenzene
1,3-Di chlorobenzene
l,4-D1chlorobenzene
Ethylbenzene
Toluene
Range
for Q
(ug/L)
15.4-24.6
16.1-23.9
13.6-26.4
14.5-25.5
13.9-26.1
12.6-27.4
15.5-24.5
L1ra1t
for s
(ug/L)
4.1
3.5
5.8
5.0
5.5
6.7
4.0
Range
for 7.
(ug/L)
10.0-27.9
12.7-25.4
10.6-27,6
12.8-25.5
11.6-25.5
10.0-28.2
11.2-27.7
Range
P. PS
(X)
39-150
55-135
37-154
50-141
42-143
32-160
46-148
      Q  =  Concentration measured  1n QC  check  sample, 1n ug/L.

      s  =  Standard  deviation  of four  recovery measurements,  1n  ug/L.

      7  =  Average recovery  for four recovery  measurements,  1n ug/L.

      P,  PS  =  Percent  recovery measured.

      aCr1ter1a are from  40 CFR  Part  136  for   Method 602 and were  calculated
 assuming  a  QC check sample concentration of  20  ug/L.  These criteria are based
 directly  upon the  method performance  data   1n   Table 4.   Where  necessary, the
 limits  for  recovery have been broadened  to  assure  applicability of  the limits
 to  concentrations  below  those used to  develop Table 1.
                                   8020 - 12
                                                          Revision      0
                                                          Date  September 1986

-------
TABLE 4.  METHOD ACCURACY AND PRECISION AS FUNCTIONS OF CONCENTRATION


Parameter
Benzene
Chlorobenzene
l,2-D1chlorobenzene
l,3-D1chlorobenzene
1 , 4-D1 ch 1 orobenzene
Ethyl benzene
Toluene
Accuracy, as
recovery, x1
(ug/L)
0.92C+0.57
0.95C+0.02
0.93C+0.52
0.96C-0.04
0.93C-0.09
0.94C+0.31
0.94C+0.65
Single analyst Overall
precision, sr'
(ug/U
0.097+0.59
0.097+0.23
0.177-0.04
0.157-0.10
0.157+0.28
0.177+0.46
0.097+0.48
precision,
S1 (ug/L)
0.217+0.56
0.177+0.10
0.227+0.53
0.197+0.09
0.207+0.41
0.267+0.23
0.187-0.71
     x1  = Expected  recovery  for  one  or  more  measurements  of  a  sample
           containing  a concentration of C, in ug/L.

     sr' = Expected  single  analyst  standard  deviation  of measurements at an
           average concentration  of 7,  in  ug/L.

     S'  = Expected  interlaboratory standard  deviation  of measurements at an
           average concentration  found  of  7,  in  ug/L.

     C   = True  value  for the  concentration,  in  ug/L.

     7   = Average recovery found for measurements of samples containing a
           concentration of C, in ug/L.
                                   8020 - 13
                                                          Revision      0
                                                          Date  September 1986

-------
                                         METHOD 6020

                                 AROMATIC VOLATILE  O«SANICS
  C    •—    )
 7. 1
 Introduce  compounds
       into g«i
   Chrom»tOOr«e»n  by
 Orect  injection or
    |M«tnoO  5030;
    7.2
o
       Set  g*c
       oimtog^i
      condition
    7.3
      Ciliortte
      (refer  tc
            603C!
7.41
  Introouc*  vol»tllt
  eomoounai  into  g*>
   cnrom»tOQr»ph  ty
    Mttnoo  5030 o-
   a>r«ct  injection
   7.4.2
          Folio-
       Section 7.6
    in Mitnoa 0000
      for «n»ly«l»
    •cauenee.  etc.
  volume puro«O
or tnjecteo «no
                           Calculate
                        concentration
                        (Section  7 . B.
                               eooo)
                       Arp  •ntlytlol
                       inter!«r«nc»»
                        •UlOCCtf07
                           Anilyit ullng
                             •econa GC
                              COlumn
                                                Dilute tecona
                                                  •1louot
       O
                                     8020 - 14
                                                                Revision       0
                                                                Date   September 1986

-------
00
o

-------
                                 METHOD 8020A

               AROMATIC VOLATILE ORGAN ICS BY GAS CHROMATOGRAPHY


1.0   SCOPE AND APPLICATION

      1.1   Method  8020  is  used  to  determine the  concentration  of  various
aromatic volatile organic compounds.   The  following compounds can be determined
by this method:
                                                   Appropriate Technique
                                                                    Direct
Compound Name                        CAS  No.a       Purge-and-Trap  Injection
Benzene
Chlorobenzene
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1, 4 -Di chlorobenzene
Ethyl benzene
Toluene
Xylenes
a Chemical Abstract
b adequate response
71-43-2
108-90-7
95-50-1
541-73-1
106-46-7
100-41-4
108-88-3

Services Registry Number.
by this technique.
b
b
b
b
b
b
b
b


b
b
b
b
b
b
b
b


      1.2   Table 1  lists  the method detection limit for each target analyte in
organic-free reagent water.  Table 2 lists the estimated quantitation limit (EQL)
for other matrices.
2.0   SUMMARY OF METHOD

      2.1   Method 8020 provides chromatographic conditions for the detection of
aromatic volatile compounds.  Samples can be introduced into the GC using direct
injection  or purge-and-trap  (Method 5030).    Ground water  samples must  be
determined  using  Method  5030.    A  temperature  program  is  used  in the  gas
chromatograph to  separate  the  organic  compounds.   Detection is  achieved  by a
photo-ionization detector (PID).

      2.2   If interferences are encountered,  the method provides  an optional gas
chromatographic column that may be helpful  in  resolving  the  analytes from the
interferences and for analyte confirmation.


3.0   INTERFERENCES

      3.1   Refer to Method 5030  and  8000.


                                   8020A  -  1                         Revision 1
                                                                September 1994

-------
      3.2   Samples  can  be  contaminated  by  diffusion  of volatile  organics
(particularly chlorofluorocarbons  and  methylene chloride)  through  the sample
container septum during shipment and storage.  A field sample blank prepared from
organic-free reagent water and carried through sampling and subsequent storage
and handling can serve as a check on such contamination.


4.0   APPARATUS AND MATERIALS

      4.1   Gas chromatograph

            4.1.1  Gas Chromatograph  -  Analytical  system  complete  with  gas
      chromatograph suitable for on-column injections or purge-and-trap sample
      introduction and all  required accessories,  including  detectors, column
      supplies, recorder,  gases,  and syringes.  A data  system for measuring peak
      heights and/or peak areas is recommended.

            4.1.2  Columns

                   4.1.2.1    Column 1:  6 ft x 0.082 in  ID #304 stainless steel
            or  glass  column packed with  5% SP-1200  and 1.75% Bentone-34  on
            100/120 mesh  Supelcoport,  or equivalent.

                   4.1.2.2    Column 2:   8 ft  x 0.1  in  ID  stainless  steel  or
            glass  column  packed  with  5%  l,2,3-Tris(2-cyanoethoxy)propane  on
            60/80 mesh Chromosorb W-AW,  or equivalent.

            4.1.3  Detector - Photoionization (PID)  (h-Nu  Systems,  Inc.  Model
      PI-51-02 or equivalent).

      4.2   Sample  introduction   apparatus  -  Refer  to  Method 5030   for  the
appropriate equipment for sample introduction purposes.

      4.3   Syringes - A 5 ml Luerlok glass  hypodermic  and a 5 ml, gas-tight with
shutoff valve.

      4.4   Volumetric flask,  Class  A  -  Appropriate  sizes with ground  glass
stoppers.

      4.5   Microsyringe  - 10  and  25 p.1 with a 0.006 in ID needle (Hamilton 702N
or equivalent) and a 100  juL.

      4.6   Analytical balance - 0.0001  g.


5.0   REAGENTS

      5.1   Organic-free  reagent water.   All references  to water in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.2   Methanol  (CH3OH) - pesticide quality or equivalent.  Store away from
other solvents.


                                   8020A -  2                         Revision 1
                                                                September 1994

-------
      5.3   Stock standards -  Stock solutions may be prepared from  pure standard
materials  or  purchased as  certified solutions.   Prepare  stock  standards in
methanol  using   assayed  liquids.    Because  of  the  toxicity  of  benzene  and
1,4-dichlorobenzene, primary dilutions of these materials  should be prepared in
a hood.

            5.3.1  Place about 9.8 ml of  methanol  in  a  10 ml tared ground glass
      stoppered volumetric flask.  Allow the  flask to  stand,  unstoppered,  for
      about 10 min or until all alcohol wetted surfaces have dried.  Weigh the
      flask to the nearest 0.0001 g.

            5.3.2  Using a 100 /xL syringe,  immediately  add  two or more drops of
      assayed reference material  to  the  flask;  then  reweigh.   The liquid must
      fall directly into the alcohol  without contacting the neck of the flask.

            5.3.3  Reweigh, dilute to volume, stopper, and  then mix by inverting
      the flask  several times.   Calculate  the concentration in  milligrams  per
      liter (mg/L) from the net gain  in weight.  When  compound purity is assayed
      to  be  96%  or  greater,  the weight  may be  used without  correction to
      calculate the concentration of  the  stock standard.  Commercially prepared
      stock standards may be used at  any  concentration  if they are certified by
      the manufacturer or by an independent source.

            5.3.4  Transfer the  stock  standard solution  into  a  Teflon-sealed
      screw-cap bottle.  Store,  with  minimal  headspace, at  4°C and protect from
      light.

            5.3.5  All  standards must be  replaced after 6  months,  or sooner if
      comparison with check standards indicates a problem.

      5.4   Secondary  dilution  standards:   Using  stock   standard  solutions,
prepare in methanol secondary dilution standards,  as needed,  that contain  the
compounds of interest,  either singly  or mixed  together.  The secondary dilution
standards should be  prepared at concentrations such that the aqueous calibration
standards  prepared  in  Section  5.5  will  bracket the working  range  of  the
analytical system.  Secondary dilution standards should be stored with  minimal
headspace for volatiles and should be  checked frequently for signs of degradation
or evaporation,  especially just  prior  to preparing calibration  standards from
them.

      5.5   Calibration standards:  Calibration standards  at a minimum  of five
concentrations  are  prepared  in  organic-free reagent water  from  the  secondary
dilution  of  the stock standards.   One of .the  concentrations  should be at  a
concentration near,  but  above,  the method  detection   limit.   The  remaining
concentrations should correspond to  the expected range  of  concentrations  found
in real samples  or  should define the working range  of  the  GC.   Each  standard
should contain each analyte for detection by this method (e.g.,  some or all of
the compounds listed in the target analyte list may be  included).   In  order to
prepare accurate aqueous standard solutions, the following precautions  must be
observed.
                                  8020A  - 3                         Revision 1
                                                                September 1994

-------
            5.5.1  Do not inject more  than  20 juL of  alcoholic  standards into
      100 ml of organic-free reagent water.

            5.5.2  Use  a 25  /iL   Hamilton   702N  microsyringe  or  equivalent
      (variations  in  needle  geometry  will  adversely affect  the  ability  to
      deliver reproducible volumes of methanolic standards into water).

            5.5.3  Rapidly  inject   the  alcoholic   standard  into  the  filled
      volumetric flask.   Remove  the needle as  fast as possible after injection.

            5.5.4  Mix aqueous standards by inverting the flask three times only.

            5.5.5  Fill  the  sample  syringe from the  standard solution contained
      in the expanded area  of the  flask  (do  not  use any  solution contained in
      the neck of the flask) .

            5.5.6  Never use pipets  to dilute or transfer  samples  or aqueous
      standards.

            5.5.7  Aqueous standards are not  stable and should be discarded after
      1 hr, unless  properly sealed and stored.   The  aqueous  standards can  be
      stored up to 24 hr, if held  in sealed  vials with zero headspace.

      5.6   Internal standards (if internal  standard calibration is  used):   To
use this approach, the analyst must select -one or more internal  standards that
are similar in  analytical behavior to  the compounds of interest.   The analyst
must further demonstrate that the  measurement of  the  internal  standard is  not
affected by method  or matrix  interferences.   Because  of  these  limitations,  no
internal  standard  can   be  suggested   that   is  applicable  to  all  samples.
Alpha, alpha, alpha-trifluorotoluene has  been  used successfully as  an  internal
standard.

            5.6.1  Prepare  calibration  standards   at  a  minimum   of  five
      concentrations for each parameter of interest as  described in Section 5.5.

            5.6.2  Prepare a spiking solution containing  each  of the  internal
      standards using the procedures described in Sections 5.3  and 5.4.  It is
      recommended  that   the secondary  dilution  standard  be  prepared  at  a
      concentration of 15 mg/L of each internal standard compound.  The addition
      of 10 /iL of this   standard to  5.0 ml  of sample  or  calibration  standard
      would be equivalent to 30
            5.6.3  Analyze each calibration standard according to Section 7.0,
      adding  10  juL  of  internal  standard  spiking  solution  directly  to  the
      syringe.

      5.7   Surrogate standards:  The analyst should monitor both the performance
of the analytical system and the effectiveness of the method in  dealing with each
sample matrix by spiking each sample, standard, and organic-free reagent water
blank with surrogate compounds (bromochlorobenzene,  bromofluorobenzene, 1,1,1-
tri fl uorotol uene,  fluorobenzene,  and difluorobenzene  are  recommended)  which
encompass the range of the temperature program  used  in this method.  From stock


                                  8020A  -  4                         Revision 1
                                                                September 1994

-------
standard solutions prepared as  in  Section  5.3,  add  a volume to give 750 /jg of
each surrogate  to 45 mi  of organic-free  reagent water contained  in  a 50 ml
volumetric flask, mix,  and  dilute to volume for a  concentration of 15 ng//zL.
Add 10  juL of this surrogate  spiking solution directly into the 5 ml  syringe with
every  sample  and reference  standard  analyzed.     If  the  internal   standard
calibration procedure is used,  the surrogate compounds  may  be  added  directly to
the internal standard spiking solution  (Section 5,6.2).


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See  the  introductory  material  to this  chapter, Organic Analytes,
Section 4.1.
7.0   PROCEDURE

      7.1   Volatile compounds are introduced into  the gas chromatograph either
by direct injection or  purge-and-trap  (Method  5030).   Method 5030 may be used
directly on ground  water samples or low-concentration  contaminated soils and
sediments.   For medium-concentration soils  or sediments, methanolic extraction,
as described in Method 5030,  may  be  necessary prior to purge-and-trap analysis.
Method 5030 also provides guidance on the analysis  of aqueous miscible and non-
aqueous miscible liquid wastes (see Section  7.4.1.1 below).

      7.2   Gas chromatography conditions (Recommended):

            7.2.1  Column 1:
            Carrier gas (He) flow rate:
            For lower boiling compounds
                   Initial  temperature:
                   Temperature program:
                       36 mL/min
                       50°C, hold  for 2
                       50°C  to 90°C at
hold until
                                              mm;
                                               6°C/mi n,
                              all  compounds  have  eluted.
For higher boiling range of compounds:
       Initial  temperature:    50°C, hold for 2 min;
       Temperature program:    50°C  to  110°C  at  3°C/min,  hold until
                                          all  compounds  have eluted
            Column  i  pruv11
      aromatic hydrocarbons
      column because of --i~
      aromatic isomers
1  provides  outstanding separations  for a wide  variety of
ocarbons.  Column 1  should  be  used  as the primary analytical
e of its unique ability to resolve para-, meta-, and ortho-
ers.
            7.2.2  Column 2:

            Carrier gas (He) flow
            Initial temperature:
            Temperature program:
               rate:    30  mL/min
                       40°C, hold  for 2 min;
                       40°C  to  100°C  at  2°C/min,  hold until
                       all  compounds  have  eluted.
                                   8020A  -  5
                                                 Revision 1
                                             September 1994

-------
            Column 2, an extremely  high  polarity column,  has been  used  for a
      number of years to resolve aromatic hydrocarbons from alkanes in complex
      samples.  However, because resolution between some of the aromatics is not
      as efficient as with  Column 1, Column 2 should be used as  a confirmatory
      column.

      7.3   Calibration:  Refer to Method 8000 for proper calibration techniques.
Use Table 1  and  especially Table 2 for guidance on selecting  the lowest point on
the calibration curve.

            7.3.1  Calibration must take place using the same  sample introduction
      method that will be  used to analyze actual samples (see Section 7.4.1).

            7.3.2  The procedure for  internal  or external  calibration may be
      used.  Refer to Method 8000 for a description of  each of these procedures.

      7.4   Gas chromatographic analysis:

            7.4.1  Introduce volatile compounds into the gas chromatograph using
      either Method 5030 (purge-and-trap method) or the direct injection method.
      If the  internal  standard calibration  technique is  used,   add 10 p,l of
      internal standard to  the sample prior to purging.

                   7.4.1.1     Direct injection:   In very  limited applications
            (e.g., aqueous  process wastes),- direct  injection of the sample into
            the  GC  system with  a  10 /iL  syringe  may be  appropriate.    The
            detection limit is very high (approximately 10,000 juQ/L); therefore,
            it is only permitted when  concentrations  in excess of  10,000  /zg/L
            are expected or for  water  soluble  compounds that do not purge.  The
            system must  be  calibrated  by direct  injection (bypassing the purge-
            and-trap device).

                   Non-aqueous miscible wastes may  also be  analyzed by direct
            injection if the  concentration of  target analytes   in  the  sample
            falls within the  calibration range.   If dilution of  the sample is
            necessary, follow the guidance  for High  Concentration  samples in
            Method 5030,  Section 7.3.3.2.

            7.4.2  Method 8000 provides instructions on the  analysis sequence,
      appropriate  dilutions,  establishing  daily  retention  time windows,  and
      identification criteria.  Include a mid-concentration standard after each
      group of 10 samples  in  the analysis sequence.

            7.4.3  Table 1 summarizes the estimated retention  times and detection
      limits for a number of organic compounds analyzable using this method.  An
      example  of  the separation achieved  by  Column  1  is  shown  in  Figure 1.
      Figure 2 shows an example of the separation achieved using  Column 2.

            7.4.4  Record the  sample volume purged or injected and the resulting
      peak sizes (in area  units or peak heights).

            7.4.5  Calculation of concentration is covered in Method 8000.
                                  8020A  - 6                         Revision 1
                                                                September 1994

-------
            7.4.6  If analytical  interferences are suspected, or for the purpose
      of confirmation, analysis using the second GC column is recommended.

            7.4.7  If the response  for  a peak is off  scale,  i.e.,  beyond the
      calibration range of the standards, prepare  a dilution of the sample with
      organic-free reagent water.   The  dilution  must  be performed on a second
      aliquot of the sample which has been properly sealed and stored prior to
      use.
8.0   QUALITY CONTROL

      8.1   Refer to Chapter  One  for specific quality  control  procedures and
Method 8000 for gas chromatographic procedures.  Quality control to ensure the
proper operation of the purge-and-trap device is covered in Method 5030.

      8.2   Quality control required to validate the GC system operation is found
in Method 8000.

            8.2.1  The  quality control check  sample  concentrate (Method 8000)
      should contain each parameter of interest  at  a concentration of  10 mg/L
      in methanol.

            8.2.2  Table 3  indicates  the calibration  and QC acceptance criteria
      for this method.   Table 4 gives method accuracy  and precision as functions
      of concentration  for the analytes  of interest.   The contents of both
      tables should be  used to evaluate a  laboratory's  ability  to perform and
      generate acceptable data by  this method.

      8.3   Calculate surrogate standard recovery on  all  samples,  blanks, and
spikes.   Determine if recovery  is within limits (limits established by performing
QC procedure outlined in Method 8000).

            8.3.1  If recovery is  not  within limits,  the following is required.

                   •    Check   to   be  sure  that  there   are   no   errors  in
                        calculations,    surrogate   solutions   and   internal
                        standards.  Also,  check instrument  performance.

                   •    Recalculate the data and/or reanalyze the extract  if any
                        of  the above checks reveal  a  problem.

                   •    Reextract  and  reanalyze the sample  if none of the above
                        are  a  problem   or   flag  the  data   as   "estimated
                        concentration".
                                  8020A  - 7                         Revision 1
                                                                September 1994

-------
9.0   METHOD PERFORMANCE

      9.1   This method was tested by 20 laboratories  using organic-free reagent
water, drinking water,  surface water, and three industrial wastewaters spiked at
six concentrations over the range 2.1  -  500  jug/L.   Single operator precision,
overall precision, and method  accuracy were found to  be directly related to the
concentration of the  parameter and essentially independent of the sample matrix.
Linear equations to describe these relationships are presented in Table 4.

      9.2   The accuracy and precision obtained will be determined  by the sample
matrix, sample introduction technique, and by the calibration procedure used.

      9.3   The method detection  limits reported in Table  1 were generated under
optimum analytical conditions  by  an  Agency  contractor (Ref.  7) as guidance, and
may not be readily achievable by all laboratories at all  times.
10.0  REFERENCES
1.


2.
3.
4.
5.


6.



7.
Bellar, T.A.
pp. 739-744,
 and J.J.
1974.
Lichtenberg,  J.  Amer.  Water Works Assoc.,  66(12),
Bellar, T.A., and J.J.  Lichtenberg,  "Semi-Automated Headspace Analysis of
Drinking  Waters  and  Industrial  Waters  for  Purgeable  Volatile  Organic
Compounds", in Van Hall  (ed.),  Measurement of Organic Pollutants in Water
and Wastewater, ASTM STP 686, pp. 108-129,  1979.

Dowty, B.J.,  S.R. Antoine, and J.L.  Laseter, "Quantitative and Qualitative
Analysis of Purgeable Organics by High Resolution Gas Chromatography and
Flame  lonization  Detection",  in Van  Hall,  ed., Measurement  of  Organic
Pollutants in Water and Wastewater.  ASTM STP 686,  pp.  24-35,  1979.

Development and Application  of Test Procedures for Specific Organic Toxic
Substances in  Wastewaters.   Category 11 - Purgeables and  Category 12 -
Acrolein,  Acrylonitrile,  and Dichlorodifluoromethane.   Report  for EPA
Contract 68-03-2635.

"EPA Method Validation Study 24, Method 602 (Purgeable Aromatics)", report
for EPA Contract 68-03-2856.

U.S. EPA 40 CFR Part 136, "Guidelines Establishing Test Procedures for the
Analysis of Pollutants Under the Clean Water  Act; Final  Rule and Interim
Final Rule and Proposed Rule", October 26,  1984.

Gebhart, J.E., S.V. Lucas, S.J. Naber, A.M.  Berry,  T.H.  Danison and H.M.
Burkholder,  "Validation of  SW-846  Methods 8010,  8015,  and  8020"; Report
for  EPA  Contract  68-03-1760,  Work  Assignment  2-15;   US  EPA,  EMSL-
Cincinnati, 1987."
                                   8020A  -  8
                                                              Revision 1
                                                          September 1994

-------
                                  TABLE 1.
            CHROMATOGRAPHIC  CONDITIONS  AND METHOD  DETECTION  LIMITS
                        FOR AROMATIC VOLATILE ORGANICS



Compound
Benzene
Chlorobenzeneb
1,4-Dichlorobenzene
1,3-Dichlorobenzene
1,2-Dichlorobenzene
Ethyl Benzene
Toluene
Xylenes
Retention
(min)

Col . 1
3.33
9.17
16.8
18.2
25.9
8.25
5.75

time


Col . 2
2.75
8.02
1-6.2
15.0
19.4
6.25
4.25

Method
detection
limit8
(M9/L)
0.2
0.2
0.3
0.4
0.4
0.2
0.2

a Using purge-and-trap method (Method 5030). See Sec.  9.3
b Chlorobenzene and m-xylene may co-elute on some columns
                                  TABLE 2.
             DETERMINATION  OF  ESTIMATED  QUANTITATION  LIMITS  (EQLs)
                             FOR VARIOUS MATRICES8
      Matrix
Factor
      Ground water
      Low-concentration soil
      Water miscible liquid waste
      High-concentration soil and sludge
      Non-water miscible waste
  10
  10
 500
1250
1250
      EQL =  [Method  detection limit  (see  Table  1)]  X  [Factor found  in  this
      table].  For non-aqueous  samples,  the factor is on  a  wet-weight basis.
      Sample EQLs are highly matrix-dependent.   The EQLs  determined herein are
      provided for guidance and may not always  be achievable.
                                  8020A  -  9
              Revision 1
          September 1994

-------
                                   TABLE 3.
                            QC ACCEPTANCE CRITERIA8


Parameter
Benzene
Chlorobenzene
1,2-Dichlorobenzene
1, 3 -Di chlorobenzene
1,4-Dichlorobenzene
Ethyl benzene
Toluene
Range
for Q
(M9/L)
15.4-24.6
16.1-23.9
13.6-26.4
14.5-25.5
13.9-26.1
12.6-27.4
15.5-24.5
Limit
for s
(M9/L)
4.1
3.5
5.8
5.0
5.5
6.7
4.0
Range
for x
(M9/L)
10.0-27.9
12.7-25.4
10.6-27.6
12.8-25.5
11.6-25.5
10.0-28.2
11.2-27.7
Range
P> PS
(%)
39-150
55-135
37-154
50-141
42-143
32-160
46-148
Q

s

x

P,

a
      Concentration measured in QC check sample,  in M9/L-

      Standard deviation of four recovery measurements,  in M9/L-

      Average recovery for four recovery measurements,  in ^g/L.

      Percent recovery measured.

Criteria from 40 CFR  Part  136  for  Method 602,  using packed columns, and
were calculated assuming a check sample concentration of 20 jug/L.  These
criteria are based directly upon the method performance data in Table 4.
Where necessary, the  limits  for recovery have been  broadened  to assure
applicability of the limits to  concentrations below those used to develop
Table 1. When capillary columns are used, see Method 8021 for performance
data.
                                  8020A - 10
                                                              Revision 1
                                                          September 1994

-------
                                   TABLE 4.
          METHOD  ACCURACY  AND  PRECISION  AS  FUNCTIONS  OF CONCENTRATION


Parameter
Benzene
Chlorobenzene
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
Ethyl benzene
Toluene
Accuracy, as
recovery, x'
(M9/L)
0.92C+0.57
0.95C+0.02
0.93C+0.52
0.96C-0.04
0.93C-0.09
0.94C+0.31
0.94C+0.65
Single analyst
precision, sf'
(M9/L)
0.09X+0.59
0.09X+0.23
0.17X-0.04
0.15X-0.10
0.15X+0.28
O.Ux+0.46
0.09X-I-0.48
Overall
precision,
S' (M9/L)
0.21X+0.56
0.17X+0.10
0.22X+0.53
0.19X+0.09
0.20X+0.41
0.26X+0.23
O.lSx+0.71
sr'


S'


C

x
Expected  recovery  for  one  or   more  measurements   of  a  sample
containing concentration C, in jug/L.

Expected  single  analyst  standard deviation of  measurements at an
average concentration of x, in
Expected interlaboratory standard  deviation  of measurements at an
average concentration found of x, in
True value for the concentration, in /zg/L.

Average  recovery  found for  measurements  of  samples  containing a
concentration of C, in jug/L.
                                  8020A - 11
                                                        Revision 1
                                                    September 1994

-------
                  Figure  1
Chromatogram of Aromatic Volatile Organics
           (column 1 conditions)
             I
Column:
Program:
Detector
Sample:
5% SP-1200/1.75% Bentone 34
50°C-2 Minutes, 6°C/Min. to 90°C
Photoionization
0.40 lig/L Standard Mixture
  U-J
                                                          i
         8      10      12      14

          RETENTION TIME (MINUTES)
                    16
                ta
20
22
                8020A - 12
                                Revision 1
                            September 1994

-------
                  Figure  2
Chromatogram of Aromatic Volatile Organics
           (column 2 conditions)
Column:
Program:
Detector:
Sample:
5% l,2,3-Tris(2-Cyanoethoxy)Propane on Chromosorb-W
40°C-2 Minutes, 2'C/Min. to lOO'C
Photoionization
2.0 ^g/L Standard Mixture
           I         12        If       30

         MffTtNTtON Tttlf (MINUTES)
                                    24
                 8020A - 13
                                       Revision 1
                                   September 1994

-------
                                     METHOD  8020A
              AROMATIC  VOLATILE ORGANICS  BY GAS CHROMATOGRAPHY
        Start
7.1  Introduce compounds
into gas chromatograph
  by direct injection or
     purge-end-trap
     (Method 5030)
      7.2 Set gas
    chromatograph
      condition.
     7.3 Calibrate
 (refer to Method SOOO)
    7.4.1 Introduce
  volatile compounds
into gas chromatograph
  by purge-and-trap or
    direct injection.
  7.4.2 Follow Method
   8000  for analysis
     sequence, etc.
7.4.4 R-ecord volume
 purged or injected
  and peak sizes.
    7.4.5 Calculate
    concentration
(refer to Method 8000)
      7.4.6 Ar«
      analytical
     interferences
     suspected?
                                             7.4.7 Is
                                           response for
                                             a peak
                                            off-scale?
                                                                    7.4.6 Analyze using
                                                                    second GC column.
                               7.4.7 Dilute second
                                 aliquot of sample.
                                        8020A  -14
                                             Revision 1
                                        September 1994

-------
00
o

-------
                                  METHOD 8021

               HALOGENATED VQLATILES BY GAS CHROMATOGRAPHY  USING
            PHOTOIONIZATIQN AND ELECTROLYTIC CONDUCTIVITY DETECTORS
                     IN  SERIES:  CAPILLARY COLUMN TECHNIQUE
1.0   SCOPE AND APPLICATION

      1.1   Method 8021  is  used to determine  volatile  organic  compounds  in a
variety of solid waste matrices.  This  method  is  applicable to nearly all types
of samples,  regardless of water content,  including ground water, aqueous sludges,
caustic  liquors,  acid  liquors,  waste  solvents,  oily wastes,  mousses,  tars,
fibrous  wastes,  polymeric  emulsions,   filter  cakes,  spent  carbons,  spent
catalysts, soils, and sediments.  The following compounds can be determined by
this method:
Analyte
CAS No.a
  Appropriate Technique
                  Direct
Purge-and-Trap    Injection
Benzene
Bromobenzene
Bromochl oromethane
Bromodichloromethane
Bromoform
Bromome thane
n-Butylbenzene
sec-Butylbenzene
tert-Butyl benzene
Carbon tetrachloride
Chlorobenzene
Chi orodi bromomethane
Chloroethane
Chloroform
Chi oromethane
2-Chlorotoluene
4-Chlorotoluene
1 , 2-Di bromo-3 -chl oropropane
1,2-Dibromoethane
Di bromomethane
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
Di chl orodi f 1 uoromethane
1,1-Dichloroethane
1,2-Dichlor'oethane
1,1-Oichloroethene
cis-l,2-Dichloroethene
trans -1,2-Oi chl oroethene
71-43-2
108-86-1
74-97-5
75-27-4
75-25-2
74-83-9
104-51-8
135-98-8
98-06-6
56-23-5
108-90-7
124-48-1
75-00-3
67-66-3
74-87-3
95-49-8
106-43-4
96-12-8
106-93-4
74-95-3
95-50-1
541-73-1
106-46-7
75-71-8
75-34-3
107-06-2
75-35-4
156-59-4
156-60-5
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
PP
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
                                   8021  -  1
                               Revision 0
                                July 1992

-------


Analyte
1,2-Dichloropropane
1,3-Dichloropropane
2,2-Oichloropropane
1, 1-Dichloropropene
cis-l,3-dichloropropene
trans-l,3-dichloropropene
Ethyl benzene
Hexachlorobutadiene
Isopropyl benzene
p-IsopropyKoluene
Methylene chloride
Naphthalene
n-Propylbenzene
Styrene
1,1,1 ,2-Tetrachloroethane
1,1,2,2-Tetrachloroethane
Tetrachloroethene
Toluene
1,2,3-Trichlorobenzene
3 ,2,4-Trichlorobenzene
1,1, 1-Trichloroethane
1,1,2-Trichloroethane
Trichloroethene
Trichlorofluoromethane
1,2,3-Trichloropropane
1,2, 4 -Tri methyl benzene
1, 3, 5 -Tri methyl benzene
Vinyl chloride
o-Xylene
m-Xylene
p-Xylene


CAS No.'
78-87-5
142-28-9
590-20-7
563-58-6
10061-01-5
10061-02-6
100-41-4
87-68-3
98-82-8
99-87-6
75-09-2
91-20-3
103-65-1
100-42-5
630-20-6
79-34-5
127-18-4
108-88-3
87-61-6
120-82-1
71-55-6
79-00-5
79-01-6
75-69-4
96-18-4
95-63-6
108-67-8
75-01-4
95-47-6
108-38-3
106-42-3
AooroDriate

Purge-and-Trap
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
D
b
b
b
b
b
b
b
Technioue
Direct
Injection
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
a Chemical Abstract Services Registry Number.
b Adequate response by thi
pp Poor purging efficiency
i Inappropriate technique
s technique.
resulting in high
for this analyte.

EQLs.




pc Poor chromatographic behavior.
       1.2   Method detection limits  (MDLs)  are compound dependent and vary with
purging  efficiency and  concentration.    The  MDLs  for  selected  analytes  are
presented in  Table 1.   The  applicable  concentration range  of  this  method is
compound  and   Instrument dependent  but  is  approximately  0.1   to 200  ng/L.'
Analytes that  are inefficiently purged  from  water will  not  be detected when
present at low concentrations,  but they can be  measured with  acceptable accuracy
and  precision when  present  in  sufficient  amounts.    Determination  of some
structural isomers (i.e. xylenes) may be hampered by  coelution.

                                   8021  - 2                         Revision 0
                                                                     July 1992

-------
       1.3   The  estimated  quantisation  limit  (EQL)  of  Method  8021   for  an
individual  compound  is  approximately  1  MQ/kg  (wet weight)  for soil/sediment
samples, 0.1 mg/kg  (wet weight)  for  wastes, and  1 ng/l  for  ground water (see
Table 3).   EQLs will be proportionately higher  for sample extracts and  samples
that require dilution or reduced sample size  to avoid saturation of the detector.

       1.4   This method  is  recommended for use only by analysts experienced in
the  measurement  of  purgeable  organics  at low  ng/L  concentrations,   or  by
experienced technicians under the close supervision of a qualified analyst.

       1.5   The toxicity or carcinogenicity of chemicals  used  in  this method has
not been  precisely  defined.   Each chemical  should be treated  as  a  potential
health  hazard,  and exposure  to these  chemicals  should  be  minimized.    Each
laboratory is responsible for maintaining awareness of OSHA  regulations regarding
safe  handling  of chemicals  used in  this  method.   Additional  references  to
laboratory  safety are available for the information of the analyst (references
4 and 6).

       1.6   The  following  method analytes have  been tentatively classified  as
known or suspected human or mammalian carcinogens: benzene,  carbon tetrachloride,
1,4-dichlorobenzene,    1,2-dichloroethane,    hexachloro-butadiene,    1,1,2,2-
tetrachloroethane,    1,1,2-trichloroethane,    chloroform,    1,2-dibromoethane,
tetrachloroethene,  trichloroethene, and  vinyl chloride. Pure standard materials
and stock standard  solutions  of  these  compounds  should be  handled in a hood.   A
NIOSH/MESA approved toxic gas respirator should be  worn when the analyst handles
high concentrations of these toxic compounds.


2.0    SUMMARY OF METHOD

       2.1   Method 8021 provides gas chromatographic conditions for the detection
of halogenated  and  aromatic volatile organic compounds.  Samples  can be analyzed
using direct injection or purge-and-trap  (Method  5030).   Ground water samples
must be analyzed using Method 5030 {where applicable).  A temperature program is
used in the gas chromatograph to separate the organic  compounds.  Detection is
achieved by  a photoionization detector  (PID)  and  an  electrolytic conductivity
detector (HECD) in series.

       2.2   Tentative Identifications  are  obtained by analyzing standards under
the same conditions  used for samples and comparing  resultant GC retention times.
Confirmatory information can be gained by comparing the  relative response from
the two detectors.   Concentrations of  the  identified components are measured by
relating the response produced for that compound to the  response produced by a
compound that is used as an internal  standard.


3.0    INTERFERENCES

       3,1   Refer  to Methods  5030 and  8000.

       3.2   Samples  can  be  contaminated  by  diffusion   of  volatile  organics
{particularly chlorofluorocarbons and methylene chloride) through  the  sample
container  septum  during shipment  and  storage.   A trip  blank  prepared  from
organic-free reagent water and carried through  sampling  and subsequent storage

                                   8021 - 3                         Revision 0
                                                                     July 1992

-------
and handling can serve as a check on such contamination.


4.0   APPARATUS  AND  MATERIALS

      4.1   Sample  introduction  apparatus  -  Refer  to  Method  5030  for  the
appropriate equipment for sample introduction purposes.

      4.2   Gas Chromatograph - capable  of temperature programming; equipped with
variable-constant  differential  flow controllers,  subambient  oven controller,
photoionization  and  electrolytic conductivity detectors connected with a short
piece of uncoated  capillary tubing, 0.32-0.5 mm  ID, and data system.

            4.2.1 Column - 60 m x  0.75 mm 10 VOCOL wide-bore capillary column
      with  1.5 jim film  thickness  (Supelco Inc.,  or equivalent).

            4.2.2  Photoionization   detector   (PID)   (Tracer   Model   703,   or
      equivalent).

            4.2.3  Electrolytic conductivity detector  (HECD) (Tracor  Hall Model
      700-A,  or  equivalent).

      4.3   Syringes - 5 ml glass  hypodermic  with Luer-Lok tips.

      4.4   Syringe valves - 2-way with Luer  ends (Teflon  or Kel-F).

      4.5   Microsyringe - 25 nL with  a 2 in. x  0.006  in.  ID,  22° bevel needle
(Hamilton #702N  or equivalent).

      4.6   Microsyringes - 10,  100 jil.

      4.7   Syringes - 0.5, 1.0,  and 5  ml, gas tight with  sliut-off valve.

      4.8   Bottles - 15 ml, Teflon lined with screw-cap or crimp top.

      4.9   Analytical balance -  0.0001 g.

      4.10  Refrigerator.

      4.11  Volumetric flasks, Class A  - 10 to 1000 ml.


5.0   REAGENTS

       5.1   Reagent grade inorganic  chemicals shall be  used  in all  tests.  Unless
otherwise indicated, it is intended that all inorganic reagents  shall  conform to
the  specifications  of  the Committee  on  Analytical  Reagents  of the American
Chemical Society,  where  such specifications are available.   Other grades may be
used, provided it  is first ascertained  that the reagent is  of sufficiently high
purity  to permit its use without  lessening the accuracy of  the determination.

       5.2   Organic-free reagent water.  All  references to  water  in this method
refer to organic-free  reagent water, as defined  in Chapter  One.


                                    8021  - 4                          Revision  0
                                                                     July 1992

-------
      5.3   Methanol,  CH3OH  -  Pesticide quality or equivalent,  demonstrated to
be free of analytes.  Store away from other solvents.

      5.4   Vinyl  chloride,  (99.9% pure), CH2=CHC1.   Vinyl chloride is available
from  Ideal  Gas Products,  Inc.,  Edison,  New  Jersey  and  from  Matheson,  East
Rutherford, New Jersey, as  well  as  from other sources.   Certified mixtures of
vinyl chloride in  nitrogen at  1.0 and 10.0  ppm (v/v)  are available from several
sources.

      5.5   Stock  standards  -  Stock  solutions may either be prepared from pure
standard materials or  purchased as certified solutions.  Prepare  stock standards
in methanol  using assayed liquids   or  gases,  as appropriate.   Because  of the
toxicity of some of the organohalides,  primary dilutions of these materials of
the toxicity should be prepared  in  a hood.

            5.5.1  Place about  9.8 ml of methanol in a 10 mL tared ground glass
      stoppered volumetric  flask.   Allow the flask to stand, unstoppered, for
      about  10 minutes until all  alcohol-wetted surfaces have dried.  Weigh the
      flask  to the  nearest  0.1 mg.

            5.5.2  Add the assayed reference material, as described below.

                  5.5.2.1      Liquids:  Using a 100 ^l syringe,  immediately add
            two or more drops  of assayed reference material  to the flask; then
            reweigh.  The liquid must  fall directly  into  the  alcohol  without
            contacting the neck of the flask.

                  5.5.2.2      Gases:    To  prepare  standards  for any compounds
            that boil below 30°C (e.g. bromomethane,  chloroethane, chloromethane,
            dichlorodifluoromethane, trichlorofluoromethane, vinyl  chloride),
            fill a 5 ml  valved gas-tight syringe with  the reference standard to
            the 5.0 ml  mark.    Lower the  needle  to  5  mm  above  the  methanol
            meniscus.  Slowly introduce  the reference standard above the surface
            of the  liquid.  The  heavy  gas rapidly  dissolves  in the methanol.
            This may also be accomplished by using a lecture bottle equipped with
            a Hamilton Lecture  Bottle Septum  (#86600).  Attach Teflon tubing to
            the side-arm relief valve and direct a gentle stream  of gas into the
            methanol meniscus.

            5.5.3  Reweigh, dilute to volume,  stopper,  and then mix by inverting
      the  flask  several  times.   Calculate the concentration in milligrams per
      liter  (mg/L)  from the net gain in  weight.  When compound purity is assayed
      to  be  96%  or  greater,  the   weight  may be  used  without  correction  to
      calculate the concentration of the .stock standard.  Commercially prepared
      stock  standards may be used at any concentration if they are certified by
      the  manufacturer  or by  an  independent  source.

            5.5.4  Transfer  the  stock standard  solution  into a  bottle  with  a
      Teflon lined  screw-cap or crimp top.  Store,  with  minimal  headspace, at -
      10°C to -20°C and protect from light.

            5.5.5  Prepare fresh  stock  standards  every  two months  for  gases.
      Reactive compounds  such as 2-chloroethyl vinyl ether and styrene may need
      to  be prepared more  frequently.   All  other  standards must  be  replaced

                                   8021  -  5                         Revision 0
                                                                     July 1992

-------
      after six months.  Both gas and liquid standards must be monitored closely
      by  comparison  to the initial  calibration curve  and by  comparison  to QC
      reference  samples.   It may be  necessary to replace  the  standards more
      frequently  if either check exceeds a 25% difference.

      5.6   Prepare secondary dilution  standards, using stock standard solutions,
in methanol,  as needed, that contain the compounds of  interest, either singly or
mixed  together.    The  secondary dilution  standards  should  be  prepared  at
concentrations such that the aqueous calibration standards prepared in Section
5.7 will  bracket  the working range of the analytical system.  Secondary dilution
standards should  be stored with  minimal  headspace for volatiles and should be
checked frequently for signs of degradation or evaporation, especially just prior
to preparing calibration standards from them.

      5.7   Calibration standards,  at a  minimum of five  concentrations  are
prepared in organic-free reagent water from the secondary dilution of the stock
standards.  One  of  the concentrations should  be  at  a concentration near,  but
above,  the  method  detection  limit.     The   remaining  concentrations  should
correspond to the expected range of  the concentrations found in  real samples or
should define  the working range  of the GC.   Standards  (one or  more)  should
contain each analyte for detection by this method  (e.g.  some or all  of the target
analytes  may  be  included).   In  order to  prepare   accurate  aqueous  standard
solutions, the following precautions must be observed.

            5.7.1  Do not  inject  more  than 20  pi of alcoholic  standards into
      100 ml of water.

            5.7.2  Use  a  25  »\.  Hamilton  702N  microsyringe  or  equivalent
      (variations in  needle  geometry will  adversely  affect  the  ability to
      deliver  reproducible volumes  of methanolic  standards  into water).

            5.7.3  Rapidly  inject  the  alcoholic  standard  into  the  filled
      volumetric  flask.  Remove the needle as  fast as possible after injection.

            5.7.4 Mix aqueous standards by inverting  the  flask three times.

            5.7.5 Fill  the sample syringe from the standard solution contained
      in  the  expanded  area of the flask (do not  use  any  solution contained in
      the neck of the  flask).

            5.7.6 Never  use  pipets  to dilute  or transfer  samples  or aqueous
      standards.

            5.7.7 Aqueous standards  are not stable and should be  discarded after
      one  hour,  unless properly sealed and stored.   The  aqueous standards can
      be  stored  up to  12 hours,  if  held  in  sealed vials  with  zero  headspace.

      5.8   Internal   standards  -  Prepare   a  spiking  solution  containing
fluorobenzene  and 2-bromo-l-chloropropane  in  methanol,  using  the procedures
described in Sections 5.5 and 5.6.  It  is recommended that the secondary dilution
standard  be  prepared  at  a concentration  of  5 mg/L  of each internal  standard
compound.   The addition  of  10  nl  of  such  a  standard to 5.0 ml  of sample or
calibration standard would be equivalent to 10 M9/L.


                                    8021 - 6                         Revision  0
                                                                     July  1992

-------
       5.9   Surrogate standards - The analyst should monitor both the performance
of the analytical  system and the effectiveness of the method in dealing with each
sample matrix by  spiking each  sample,  standard, and reagent blank with two or
more  surrogate  compounds.    A  combination of  bromochloromethane,  2-bromo-l-
chloropropane and  1,4-dichlorobutane is recommended to encompass the range of the
temperature program used  in this method.  From stock standard solutions prepared
as in  Section 5.5,  add a volume to give 750  fig of each surrogate to 45 ml of
organic-free reagent water contained  in a 50 ml volumetric flask, mix, and dilute
to volume for a  concentration of 15  ng/juL.   Add  10  nl of this  surrogate spiking
solution directly into the 5 ml syringe with every sample and reference standard
analyzed.  If the  internal standard calibration procedure is used,  the surrogate
compounds  may be  added  directly  to the  internal  standard  spiking  solution
(Section 5.8).


6.0    SAMPLE  COLLECTION, PRESERVATION,  AND HANDLING

       6.1   See  the  introductory material  to this  chapter,  Organic Analytes,
Section 4.1.
7.0    PROCEDURE

      7.1   Volatile  compounds  are  introduced  into the gas chromatograph either
by direct injection  or purge-and-trap  (Method 5030).   Method  5030 may be used
directly on ground water samples or low-concentration  contaminated  soils and
sediments.   For medium-concentration  soils  or  sediments, methanolic extraction,
as described in Method 5030,  may  be necessary  prior to purge-and-trap analysis.
Method 5030 also provides guidance  on the analysis of aqueous miscible and non-
aqueous miscible liquid wastes (see Section 7.4.1.1 below).

      7.2   Gas chromatography  conditions  (Recommended)

            7.2.1  Set  up   the   gas   chromatograph   system   so   that   the
      photoionization  detector  (PID)   is  in  series  with  the  electrolytic
      conductivity detector  (HECD).

            7.2.2  Oven settings:

            Carrier gas (Helium)  Flow rate:     6 mL/min.
            Temperature program
                  Initial temperature:           10°C,  hold for 8 minutes
                  Program:                       10°C  to 180°C  at  4°C/min
                  Final  temperature:             180°C,  hold until all expected
                                                compounds have eluted.

            7.2.3  The carrier gas flow  is  augmented with  an additional 24 ml of
      helium  flow before entering  the  photoionization  detector.  This make-up
      gas is  necessary to ensure optimal response from both detectors.

            7.2.4  These halogen-specific  systems  eliminate misidentifications
      due to  non-organohalides which are coextracted  during the purge  step.  A
                                   8021 - 7                         Revision 0
                                                                     July 1992

-------
      Tracer Hall Model 700-A detector was  used  to gather the single laboratory
      accuracy  and  precision  data   presented  in  Table  2.    The  operating
      conditions  used  to collect these data are:

            Reactor tube:                        Nickel, 1/16 in 00
            Reactor temperature:                 810°C
            Reactor base  temperature:            250°C
            Electrolyte:                         100% n-Propyl  alcohol
            Electrolyte flow  rate:               0.8 mL/min
            Reaction gas:                        Hydrogen at 40 mL/min
            Carrier gas plus  make-up gas:        Helium at 30 mL/min

            7.2.5  A sample chromatogram obtained with this column is presented
      in  Figure 5.   This  column was  used to develop the  method performance
      statements  in Section  9.0.   Estimated  retention times and MOLs that can
      be  achieved under  these conditions  are  given  in Table 1.  Other columns
      or  element  specific detectors may be used if the requirements of Section
      8.0 are met.

      7.3  Calibration - Refer to Method 8000 for proper calibration techniques.
Use Table 1  and  especially Table 2 for guidance on selecting the lowest point on
the calibration curve.

            7.3.1  Calibration must take place using the same sample introduction
      method that will be used to analyze  actual samples (see Section 7.4.1).

            7.3.2  The  procedure  for  internal  or external  calibration may be
      used.  Refer to Method  8000 for a description of each of these procedures.

      7.4  Gas  chromatographic  analysis

            7.4.1  Introduce volatile compounds into the gas chromatograph using
      either Method 5030 (purge-and-trap method) or the direct injection method
      (see  Section 7.4.1.1).  If the internal  standard  calibration technique is
      used, add 10  juL of internal  standard to the sample prior  to purging.

                  7.4.1.1      Direct  injection  -  In  very limited applications
            (e.g.  aqueous process wastes) direct  injection of the sample into the
            GC  system with a  10 nl syringe may  be  appropriate.   The detection
            limit  is very high  (approximately 10,000 M9/L),  therefore,  it is
            only permitted where concentrations  in  excess  of  10,000 ng/l are
            expected or  for  water-soluble  compounds  that  do  not purge.   The
            system must be calibrated  by direct  injection (bypassing the purge-
            and-trap device).

                  Non-aqueous miscible wastes may  also be  analyzed  by  direct
            injection if the concentration of target analytes in the sample falls
            within  the calibration  range.    If  dilution  of  the   sample  is
            necessary,  follow the  guidance for High  Concentration  samples in
            Method 5030,  Section 7.3.3.2.

            7.4.2  Follow  in  Method   8000  for  instructions  on  the  analysis
      sequence, appropriate  dilutions, establishing  daily  retention  time


                                   8021 -  8                         Revision 0
                                                                     July 1992

-------
      windows,  and  identification  criteria.    Include  a  mid-concentration
      standard after each group of 10 samples in the analysis sequence.

            7.4.3  Table 1 summarizes the estimated retention  times  on  the two
      detectors for a number of organic  compounds analyzable using this method.

            7.4.4  Record the sample volume purged or injected and the resulting
      peak  sizes  (in area units or peak heights).

            7.4.5  Calculation of concentration  is covered in  Method  8000.

            7.4.6  If analytical  interferences are suspected, or for the purpose
      of confirmation,  analysis using a second GC column  is recommended.

            7.4.7  If the response  for a peak is  off-scale,  prepare a dilution of
      the  sample  with  organic-free  reagent  water.    The  dilution  must  be
      performed on a second  aliquot of the sample which  has been properly sealed
      and stored  prior  to use.


8.0   QUALITY CONTROL

      8.1   Refer  to Chapter One  for specific quality  control  procedures, and
Method 8000 for gas chromatographic procedures.   Quality control  to  ensure the
proper operation of the purge-and-trap device is covered in Method 5030.

      8.2   Quality control required to validate the GC system operation is found
in Method 8000, Section 8.6.

            8.2.1  The quality control reference sample (Method 8000, Step 8.6)
      should contain each parameter  of  interest  at  a concentration  of 10 mg/L
      in methanol.

            8.2.2  Table 2 gives method  accuracy  and  precision as  functions of
      concentration  for the  analytes of interest.

      8.3   Calculate surrogate standard recovery on all  samples,  blanks, and
spikes.   Determine if recovery is within limits (limits established by performing
QC procedure outlined in Method 8000, Section 8.10).

      If recovery is not within limits, the following  is  required.

      •     Check  to be sure there  are  no  errors in  calculations,  surrogate
            solutions and internal  standards.  Also check instrument performance.

      •     Recalculate the data and/or reanalyze  the extract if any of the  above
            checks reveal a  problem.

      •     Re-extract   and  re-analyze  the  sample if none of the above are a
            problem or  flag  the data  as  "estimated  concentration".
                                   8021 - 9                         Revision 0
                                                                     July 1992

-------
9.0   METHOD PERFORMANCE

      9.1   Method detection limits for these analytes have been calculated from
data collected by spiking organic-free  reagent water  at 0.1  ng/L.   These data
are presented in Table 1.

      9.2   This  method was tested  in  a sinjgle  laboratory using organic-free
reagent water spiked at 10 jigA-   Single laboratory precision and accuracy data
for each detector are presented for the method analytes in Table 2.


10.0  REFERENCES

1.    Volatile Organic Compounds  in Water by Purqe-and-Trap Capillary Column Gas
      Chromatoaraphv   with  Photoionization   and   Electrolytic  Conductivity
      Detectors in Series. Method 502.2; U.S. Environmental Protection Agency.
      Environmental   Monitoring   and   Support   Laboratory:   Cincinnati,   OH,
      September,  1986.

2.    The Determination of Haloqenated Chemicals in Water  bv  the  Purge and Trap
      Method.  Method  502.1;  Environmental  Protection Agency,  Environmental
      Monitoring  and Support  Laboratory:  Cincinnati, Ohio  45268,  September,
      1986.

3.    Volatile Aromatic and Unsaturated Organic Compounds  in  Water by Purge and
      Trap  Gas  Chromatograohv. Method  503.1;  Environmental Protection Agency,
      Environmental   Monitoring   and   Support  Laboratory:  Cincinnati,  Ohio,
      September,  1986.

4.    Glaser, J.A.;  Forest, D.L.; McKee, G.D.; Quave,  S.A.; Budde, W.L. "Trace
      Analyses  for Wastewaters";  Environ.  Sci. Techno!. 1981,  15,  1426.

5.    Bellar,  T.A.;  Lichtenberg, J.J.  The  Determination  of Synthetic Organic
      Compounds in Water  by Purge and  Sequential Trapping  Capillary Column Gas
      Chromatographv;   U.S.  Environmental  Protection Agency,  Environmental
      Monitoring  and Support Laboratory: Cincinnati,  Ohio, 45268.
                                   8021 - 10                         Revision  0
                                                                     July  1992

-------
                              TABLE 1.

CHROMATOGRAPHIC RETENTION TIMES AND METHOD DETECTION LIMITS (MDL)  FOR
  VOLATILE ORGANIC COMPOUNDS ON PHOTOION IZATION DETECTION (PID) AND
      HALL  ELECTROLYTIC  CONDUCTIVITY  DETECTOR  (HECD) DETECTORS
Analyte
Di chl orodi f 1 uoromethane
Chloromethane
Vinyl Chloride
Bromomethane
Chloroethane
Tri chl orof 1 uoromethane
1,1-Dichloroethene
Methyl ene Chloride
trans-1 ,2-Dichloroethene
1,1-Dichloroethane
2,2-Dichloropropane
cis-1, 2 -Di chl oroethane
Chloroform
Bromochl oromethane
1,1,1 -Trichl oroethane
1,1-Dichloropropene
Carbon Tetrachloride
Benzene
1,2-Dichloroethane
Trichloroethene
1,2-Dichloropropane
Bromodi chl oromethane
Dibromomethane
Toluene
1 , 1 , 2-Trichl oroethane
Tetrachloroethene
1,3-Dichloropropane
Di bromochl oromethane
1,2-Dibromoethane
Chlorobenzene
Ethyl benzene
1,1,1, 2 -Tetrachl oroethane
m-Xylene
p-Xylene
o-Xylene
Styrene
Isopropyl benzene
Bromoform
1,1, 2, 2 -Tetrachl oroethane
1 , 2 , 3-Trichl oropropane


PID
Ret. Time8
minute
_b
-
9.88
-
_
_
16.14
-
19.30
-
-
23.11
-
.
-
25.21
-
26.10
-
27.99
-
-
_
31.95
-
33.88
-
-
-
36.56
36.72
-
36.98
36.98
38.39
38.57
39.58
-
-
-
8021 - 11

HECD
Ret. Time
minute
8.47
9.47
9.93
11.95
12.37
13.49
16.18
18.39
19.33
20.99
22.88
23.14
23.64
24.16
24.77
25.24
25.47
-
26.27
28.02
28.66
29.43
29.59
-
33.21
33.90
34.00
34.73
35.34
36.59
-
36.80
-
-
-
-
-
39.75
40.35
40.81


PID
MDL
M9A


0.02



NDC

0.05


0.02



0.02

0.009

0.02



0.01

0.05



0.003
0.005

0.01
0.01
0.02
0.01
0.05





HECD
MDL
M9A
0.05
0.03
0.04
1.1
0.1
0.03
0.07
0.02
0.06
0.07
0.05
0.01
0.02
0.01
0.03
0.02
0.01

0.03
0.01
0.006
0.02
2.2

ND
0.04
0.03
0.03
0.8
0.01

0.005





1.6
0.01
0.4
Revision 0
July 1992

-------
                                   TABLE 1.
                                  (Continued)
Analyte
   PIO
Ret. Time8
  minute
  HECO
Ret. Time
 minute
PID
MDL
M9/L
HECO
 MDL
M9/L
n-Propylbenzene                   40.87
Bromobenzene                      40.99
1,3,5-Trimethylbenzene            41.41
2-Chlorotoluene                   41.41
4-Chlorotoluene                   41.60
tert-Butylbenzene                 42.92
1,2,4-Trimethy]benzene            42.71
sec-Butyl benzene                  43.31
p-Isopropyltoluene                43.81
1,3-Dichlorobenzene               44.08
1,4-Dichlorobenzene               44.43
n-Butyl benzene                    45.20
1,2-Dichlorobenzene               45.71
l,2-Dibromo-3-Chloropropane
1,2,4-Trichlorobenzene            51.43
Hexachlorobutadiene               51.92
Naphthalene                       52.38
1,2,3-Trichlorobenzene            53.34

Internal Standards
  Fluorobenzene                   26.84
  2-Bromo-1-chloropropane
                 41.03

                 41.45
                 41.63
                 44.11
                 44.47

                 45.74
                 48.57
                 51.46
                 51.96

                 53.37
                 33.08
                             0.
                             0.
                             0.
             0.004
             0.006
             0.004
             NO
             0.02
               .06
               ,05
               .02
             0.01
             0.02
             0.007
             0.02
             0.05

             0.02
             0.06
             0.06
             ND
          0.03

          0.01
          0.01
          0.02
          0.01
          0.02
          3.0
           .03
           .02
0.
0.
          0.03
a   Retention times determined  on  60 m x  0.75 mm ID  VOCOL  capillary column.
    Program:  Hold at 10°C for 8 minutes, then program  at 4°C/min to 180°C, and
    hold until  all  expected compounds have  eluted.

b   Dash (-)  indicates  detector does not respond.

c   ND = Not  determined.
                                   8021 - 12
                                   Revision 0
                                    July 1992

-------
                   TABLE1 2.
SINGLE LABORATORY ACCURACY AND PRECISION DATA
   FOR VOLATILE ORGANIC COMPOUNDS IN WATERd
Photoionization
Detector
Analyte
Benzene
Bromobenzene
Bromochl oromethane
Bromodichloromethane
Bromoform
Bromomethane
n-Butyl benzene
sec-Butyl benzene
tert-Butylbenzene
Carbon tetrachloride
Chlorobenzene
Chloroethane
Chloroform
Chi oromethane
2-Chlorotoluene
4-Chlorotoluene
l,2-Dibromo-3-chloropropane
Di bromochl oromethane
1,2-Dibromoethane
Dibromomethane
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
Dichl orodi f 1 uoromethane
1,1-Dichloroethane
1,2-Dichloroethane
1,1-Dichloroethene
cis-1,2 Dichloroethene
trans -1,2-Dichloroethene
1,2-Dichloropropane
1,3-Dichloropropane
2,2-Dichloropropane
1,1-Dichloropropene
Ethyl benzene
Hexachlorobutadiene
I sopropyl benzene
p- Isopropyl tol uene
Recovery,8
%
99
99
-
-
-
-
100
97
98
-
100
-
.
-
NDC
101
-
-
-
.
102
104
103
-
-
-
100
ND
93
-
-
-
103
101
99
98
98
Standard
Deviation
of Recovery
1.2
1.7
-
-
-
_
4.4
2.6
2.3
-
1.0
-
_
-
ND
1.0
-
-
-
-
2.1
1.7
2.2
-
-
-
2.4
ND
3.7
-
-
-
3.6
1.4
9.5
0.9
2.4
Hall Electrolytic
Conductivity Detector
Standard
Recovery,8 Deviation
% of Recovery
_b
97
96
97
106
97
-
-
-
92
103
96
98
96
97
97
86
102
97
109
100
106
98
89
100
100
103
105
99
103
100
105
103
-
98
-
-

2.7
3.0
2.9
5.5
3.7
-
-
-
3.3
3.7
3.8
2.5
8.9
2.6
3.1
9.9
3.3
2.7
7.4
1.5
4.3
2.3
5.9
5.7
3.8
2.9
3.5
3.7
3.8
3.4
3.6
3.4
-
8.3
-
-
                   8021 - 13
Revision 0
 July 1992

-------
                                         TABLE 2.
                                       (Continued)
Analyte
Photoionizat
Detector
Recovery,"
%
ion
Standard
Deviation
of Recovery
Hall Electrolytic
Conductivity Detector
Standard
Recovery,8 Deviation
% of Recovery
Methyl ene chloride
Naphthalene
n-Propyl benzene
Styrene
1,1, 1 ,2-Tetrachl oroethane
1,1,2 , 2-Tetrachl oroethane
Tetrachloroethene
Toluene
1,2,3-Trichlorobenzene
1,2,4-Trichlorobenzene
1 , 1 , 1-Trichl oroethane
1,1,2-Trichl oroethane
Trichloroethene
Trichlorofluoromethane
1,2,3-Trichloropropane
1, 2, 4-Trimethyl benzene
1,3, 5- Trimethyl benzene
Vinyl chloride
o-Xylene
m-Xylene
p-Xylene
-
102
103
104
-
-
101
99
106
104
-
-
100
-
-
99
101
109
99
100
99
-
6.3
2.0
1.4
-
-
1.8
0.8
1.9
2.2
-
-
0.78
-
-
1.2
1.4
5.4
0.8
1.4
0.9
97
-
-
-
99
99
97
-
98
102
104
109
96
96
99
-
-
95
-
-
~
2.8
-
-
-
2.3
6.8
2.4
.
3.1
2.1
3.4
6.2
3.5
3.4
2.3
-
-
5.6

-
-
a   Recoveries  and standard deviations  were determined from  seven  samples and  spiked  at
    10  M9/L of each analyte. Recoveries were determined by internal standard method. Internal
    standards were: Fluorobenzene  for PID, 2-Bromo-l-chloropropane for HECD.

b   Detector does  not  respond.

c   ND  =  Not determined.
    This  method was  tested
    reference  8).
in a  single laboratory  using  water  spiked  at  10 ^g/L (see
                                         8021  -  14
                                             Revision 0
                                              July 1992

-------
                      TABLE 3.
DETERMINATION OF ESTIMATED QUANTITATION LIMITS (EQL)
                FOR VARIOUS MATRICES"
  Matrix                              Factor6
  Ground water                             10
  Low-concentration soil                   10
  Water miscible liquid waste             500
  High-concentration soil and sludge     1250
  Non-water miscible waste               1250
  Sample EQLs are highly matrix dependent.  The EQLs listed herein
  are provided for guidance and may not always be achievable.

  EQL =  [Method  detection  limit (Table  1)]  X [Factor (Table 2)].
  For non-aqueous samples, the factor is on a wet-weight basis.
                      8021 - 15                         Revision  0
                                                        July  1992

-------
                          FIGURE 1.
                       PURGING DEVICE
                         . on «• • oo
                          FIGURE 2.
TRAP PACKINGS AND  CONSTRUCTION TO INCLUDE DESORB CAPABILITY
                »«CXMO OVTA*.
                          8021 - 16
Revision 0
 July 1992

-------
                     FIGURE  3.
       PURGE-AND-TRAP  SYSTEM - PURGE MODE
  CAMWMOA*
                                  NOT*
                                  *U. LJMS MTVMCN TH**
                                  *MO QC SMOMO M HCATV
                                  rowx.
                     FIGURE  4.
SCHEMATIC OF PURGE-AND-TRAP DEVICE  -  DESORB  MODE
                                   HOTt
                                   *u. UNCS WTWUN nv*
                                   AM3 OC V4XLD M >**!»
                     8021  -  17
Revision  0
 July  1992

-------
COLUMNI  6O  METER  x O.73  MM  I.D.  VOCOL CAPILLARY


(•UftCC AMD THAT  VOC'm WITH HALL ft  PIO IN SERIES

     •*m i*n   no «   f»*5  l^-> —   B lOMHi—araiQ  m«« a    «a
     MN iw>   »— H   Ka  »- o   E iii»-M<5-«n a  ONO) <3    »•*
                                                                                   Snux  rs n   O-. (ucrnr^  r-.
                                                                                   MOQ  Kl »!   ^ (X\JOJI-»  •
00
o
rv>
oo
                                                                                                                  o
                                                                                                                  5>
                                                                                                                  to
                                                                                                  yuL PID
                                                                                                                  •n 30
                                                                                                                    m

                                                                                                                  O en
                                                                                                                  O
                                                                                                                  30
                                                                                                                  CD
                                                                                                            HECD

-------
                                METHOD 8021
         HALOGENATED VOLATILES BY  GAS CHROMATOGRAPHY  USING
PHOTOIONIZATION  AND ELECTROLYTIC  CONDUCTIVITY  DETECTORS  IN SERIES:
                       CAPILLARY COLUMN TECHNIQUE
            7 3 R.f.r to
            Method 8000
                for
            calibration
            t«chniqu««.
741 Introduce
aa»ple into CC
 \»ing direct
 injection or
purge-and-trip
            744 Record
            t*mpl» foluam
            introduced
            into CC and
                                 7 IS R.f.r
                                  to Method
                                  8000 for
                                caIculat
                       746 An
                      analytical
                     int«rf«c«nc«»
                      •uap«et«d?
• acapla
 >«eond CC
  coluan
                                                      Dilul* and
                                                       r«analy>>
                                                        ••cond
                                                      aliquot of
                                  8021 -  19
                                                               Revision 0
                                                                July 1992

-------
00
©

-------
                                 METHOD 8021A

               HAL06ENATED  VOLATILES  BY  GAS  CHROMATOGRAPHY  USING
            PHOTOIONIZATION AND  ELECTROLYTIC CONDUCTIVITY DETECTORS
                     IN  SERIES:  CAPILLARY COLUMN TECHNIQUE
1.0   SCOPE AND APPLICATION

      1.1   Method 8021  is  used  to determine volatile organic  compounds  in  a
variety of solid waste matrices.  This method  is applicable to nearly all types
of samples, regardless of water content, including ground water, aqueous sludges,
caustic  liquors,  acid liquors,  waste  solvents,  oily  wastes,
fibrous  wastes,  polymeric  emulsions,  filter  cakes,  spent
catalysts, soils, and sediments.   The following compounds can
this method:
                           mousses,  tars,
                          carbons,  spent
                         be determined by
Analyte
                                                  Appropriate Technique
CAS No.a   Purge-and-Trap
Direct
Injection
Benzene
Bromobenzene
Bromochl oromethane
Bromodichloromethane
Bromoform
Bromomethane
n-Butyl benzene
sec-Butylbenzene
tert-Butyl benzene
Carbon tetrachloride
Chlorobenzene
Chlorodibromomethane
Chloroethane
Chloroform
Chloromethane
2-Chlorotoluene
4-Chlorotoluene
l,2-Dibromo-3-chloropropane
1,2-Dibromoethane
Dibromomethane
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
Dichlorodifluoromethane
1,1-Dichloroethane
1,2-Dichloroethane
1,1-Dichloroethene
cis-l,2-Dichloroethene
trans-l,2-Dichloroethene
71-43-2
108-86-1
74-97-5
75-27-4
75-25-2
74-83-9
104-51-8
135-98-8
98-06-6
56-23-5
108-90-7
124-48-1
75-00-3
67-66-3
74-87-3
95-49-8
106-43-4
96-12-8
106-93-4
74-95-3
95-50-1
541-73-1
106-46-7
75-71-8
75-34-3
107-06-2
75-35-4
156-59-4
156-60-5
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
PP
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
                                  8021A  - 1
                               Revision 1
                           September 1994

-------
Analyte
CAS No.'
  Appropriate Technique
                  Direct
Purge-and-Trap    Injection
1,2-Dichloropropane
1,3-Dichloropropane
2,2-Dichloropropane
1,1-Dichloropropene
cis-l,3-dichloropropene
trans-1 ,3-dichloropropene
Ethyl benzene
Hexachlorobutadiene
Isopropyl benzene
p-Isopropyl toluene
Methylene chloride
Naphthalene
n-Propylbenzene
Styrene
1,1,1, 2 -Tetrachl oroethane
1,1,2,2-Tetrachloroethane
Tetrachloroethene
Tol uene
1,2,3-Trichlorobenzene
1,2,4-Trichlorobenzene
1,1,1-Trichloroethane
1,1,2-Trichloroethane
Trichloroethene
Tri chl orof 1 uoromethane
1,2,3-Trichloropropane
1, 2, 4-Tri methyl benzene
1,3, 5 -Tri methyl benzene
Vinyl chloride
o-Xylene
m-Xylene
p-Xylene
78-87-5
142-28-9
590-20-7
563-58-6
10061-01-5
10061-02-6
100-41-4
87-68-3
98-82-8
99-87-6
75-09-2
91-20-3
103-65-1
100-42-5
630-20-6
79-34-5
127-18-4
108-88-3
87-61-6
120-82-1
71-55-6
79-00-5
79-01-6
75-69-4
96-18-4
95-63-6
108-67-8
75-01-4
95-47-6
108-38-3
106-42-3
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
a Chemical Abstract Services Registry Number.
b Adequate response by thi
pp Poor purging efficiency
s technique.
resulting in high EQLs.




      1.2   Method detection limits (MDLs) are compound dependent  and vary with
purging  efficiency  and  concentration.    The  MDLs  for selected  analytes  are
presented  in  Table  1.   The  applicable  concentration range of  this  method is
compound  and   instrument dependent  but  is   approximately  0.1  to  200  p.g/1,
Analytes that  are  inefficiently purged  from  water will not  be detected when
present at low concentrations, but they can be  measured  with acceptable accuracy
and  precision when  present  in  sufficient amounts.    Determination of some
structural isomers (i.e. xylenes) may be hampered by coelution.
                                   8021A -  2
                               Revision 1
                           September 1994

-------
      1.3   The  estimated  quantitation  limit  (EQL)  of  Method  8021A  for  an
individual compound  is  approximately 1  jug/kg  (wet weight)  for  soil/sediment
samples, 0.1 mg/kg  (wet  weight)  for wastes, and 1 p.g/1  for ground water (see
Table 3).  EQLs will be proportionately higher  for sample extracts and samples
that require dilution or reduced sample size  to avoid saturation of the detector.

      1.4   This method  is  recommended for  use  only by analysts experienced in
the measurement of purgeable organics at the  low p-g/L level or by experienced
technicians under the close supervision of  a qualified analyst.

      1.5   The toxicity or carcinogenicity  of chemicals used in this method has
not been  precisely  defined.  Each  chemical  should be treated as  a  potential
health  hazard,  and exposure  to these  chemicals   should  be minimized.   Each
laboratory is responsible for maintaining  awareness of OSHA regulations regarding
safe  handling  of chemicals used  in  this   method.   Additional  references  to
laboratory safety are available for the information of the analyst (references
4 and 6).

      1.6   The following  method analytes  have  been  tentatively  classified as
known or suspected human or mammalian carcinogens: benzene,  carbon tetrachloride,
1,4-dichlorobenzene,    1,2-dichloroethane,    hexachloro-butadiene,    1,1,2,2-
tetrachloroethane,    1,1,2-trichloroethane,    chloroform,   1,2-dibromoethane,
tetrachloroethene,  trichloroethene, and vinyl chloride. Pure standard materials
and stock standard  solutions of  these  compounds  should be  handled in a hood.  A
NIOSH/MESA approved toxic gas respirator should be worn when  the analyst handles
high concentrations of these toxic  compounds.


2.0   SUMMARY OF METHOD

      2.1   Method  8021   provides  gas  chromatographic  conditions  for  the
detection of halogenated and aromatic volatile organic compounds.  Samples can
be analyzed using direct injection or purge-and-trap (Method 5030). Ground water
samples must be analyzed  using Method 5030  (where  applicable).   A temperature
program  is used  in the  gas chromatograph  to separate the  organic  compounds.
Detection is achieved by  a photoionization  detector  (PID) and an electrolytic
conductivity detector (HECO) in series.

      2.2   Tentative identifications are obtained  by  analyzing standards under
the same conditions  used  for samples and  comparing resultant  GC retention times.
Confirmatory information can be gained by comparing the relative  response from
the two detectors.   Concentrations  of  the identified components are measured by
relating the response produced for  that  compound to the response  produced by a
compound that is used as an internal standard.


3.0   INTERFERENCES

      3.1   Refer to Methods 5030  and 8000.

      3.2   Samples  can  be contaminated  by diffusion  of volatile  organics
(particularly chlorofluorocarbons  and methylene chloride)  through the  sample
container  septum  during  shipment  and storage.   A trip  blank  prepared  from


                                   8021A  - 3                        Revision 1
                                                                September 1994

-------
organic-free reagent water and carried through sampling and subsequent storage
and handling can serve as a check on such contamination.

      3.3   Sulfur dioxide is a potential  interferant  in the analysis for vinyl
chloride.
4.0   APPARATUS AND MATERIALS

      4.1    Sample  introduction  apparatus  -  Refer  to  Method  5030 for  the
appropriate equipment for sample introduction purposes.

      4.2    Gas  Chromatograph  - capable of  temperature  programming; equipped
with variable-constant differential flow controllers, subambient oven controller,
photoionization and electrolytic conductivity detectors connected with a short
piece of uncoated capillary tubing, 0.32-0.5 mm ID, and data system.

             4.2.1  Column  -  60  m x 0.75 mm ID VOCOL wide-bore capillary column
      with 1.5 fj.m  film thickness  (Supelco Inc., or  equivalent).

             4.2.2  Photoionization   detector  (PID)   (Tracer   Model   703,   or
      equivalent).

             4.2.3  Electrolytic conductivity detector  (HECD)  (Tracor Hall Model
      700-A, or equivalent).

      4.3    Syringes - 5 ml glass  hypodermic with  Luer-Lok tips.

      4.4    Syringe valves  - 2-way with  Luer ends  (Teflon or Kel-F).

      4.5    Microsyringe -  25  fj,l  with  a  2 in. x 0.006  in. ID,  22° bevel  needle
(Hamilton #702N or  equivalent).

      4.6    Microsyringes  - 10, 100 /iL.

      4.7    Syringes - 0.5, 1.0,  and 5 ml, gas-tight with shut-off  valve.

      4.8    Bottles - 15 ml, Teflon lined with screw-cap or  crimp top.

      4.9    Analytical balance  - 0.0001  g.

      4.10   Refrigerator.

      4.11   Volumetric flasks,  Class  A  - Appropriate  sizes  with ground glass
stoppers.


5.0   REAGENTS

      5.1    Reagent  grade  inorganic  chemicals  shall  be used  in  all  tests.
Unless otherwise indicated,  it is intended  that all  inorganic reagents shall
conform to the  specifications  of  the  Committee on  Analytical  Reagents  of the
American Chemical Society, where such specifications  are available. Other grades


                                   8021A  - 4                         Revision 1
                                                                September 1994

-------
may be used, provided it is first  ascertained that the reagent is of sufficiently
high  purity  to  permit  its  use  without  lessening  the  accuracy  of  the
determination.

      5.2   Organic-free  reagent water.  All  references to  water  in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.3   Methanol, CH3OH - Pesticide quality or equivalent, demonstrated to
be free of analytes.  Store away from other solvents.

      5.4   Vinyl chloride, (99.9% pure), CH2=CHC1.  Vinyl chloride is available
from  Ideal  Gas  Products,  Inc.,  Edison,  New Jersey  and  from Matheson,  East
Rutherford, New Jersey, as well  as  from  other  sources.  Certified  mixtures of
vinyl chloride in nitrogen at  1.0 and  10.0  ppm  (v/v)  are available from several
sources.

      5.5   Stock standards - Stock solutions may either be prepared from pure
standard materials or purchased as certified solutions.  Prepare stock standards
in methanol using  assayed liquids or  gases,  as appropriate.  Because  of the
toxicity of some of the organohalides, primary dilutions of these materials of
the toxicity should be prepared in a hood.

      NOTE: If  direct  injection  is  used,  the  solvent system of standards must
            match  that  of the  sample.   It  is  not  necessary to prepare high
            concentration aqueous mixed standards when using direct injection.

            5.5.1  Place  about 9.8 ml  of methanol in  a  10 ml tared ground glass
      stoppered  volumetric flask.   Allow the  flask  to stand, unstoppered, for
      about 10 minutes until  all  alcohol-wetted surfaces have dried.  Weigh the
      flask to  the  nearest 0.1 mg.

            5.5.2 Add  the assayed  reference material,  as described  below.

                  5.5.2.1    Liquids: Using a  100 /A  syringe, immediately add
            two or  more drops of assayed reference material to the flask; then
            reweigh.   The liquid must fall directly  into  the  alcohol  without
            contacting the neck  of  the flask.

                  5.5.2.2    Gases:   To  prepare standards  for  any compounds
            that    boil   below   30°C   (e.g.    bromomethane,    chloroethane,
            chloromethane,   dichlorodifluoromethane,   trichlorofluoromethane,
            vinyl  chloride),  fill  a  5 mL valved  gas-tight syringe with the
            reference  standard  to  the 5.0  ml mark.    Lower  the needle  to 5 mm
            above  the  methanol  meniscus.    Slowly  introduce  the  reference
            standard above the  surface of the  liquid.  The  heavy  gas  rapidly
            dissolves in  the methanol.  This may also be accomplished by using
            a  lecture  bottle equipped with  a  Hamilton Lecture  Bottle  Septum
            (#86600).   Attach  Teflon  tubing  to the  side-arm  relief valve and
            direct  a gentle stream of  gas  into  the methanol meniscus.

            5.5.3 Reweigh, dilute to volume,  stopper, and then mix by inverting
      the  flask several times.   Calculate  the  concentration in  milligrams per
      liter (mg/L) from the net gain in weight.   When  compound purity is assayed


                                  8021A  -  5                         Revision 1
                                                                September 1994

-------
      to  be  96%  or  greater,  the  weight  may  be  used  without  correction  to
      calculate the concentration of the stock  standard.  Commercially prepared
      stock standards may be used at any concentration if they are certified by
      the manufacturer or by an  independent source.

            5.5.4  Transfer  the  stock  standard  solution  into a  bottle  with a
      Teflon lined screw-cap or  crimp  top.   Store,  with  minimal  headspace,  at
      -10°C to -20°C  and  protect  from light.

            5.5.5  Prepare fresh  stock  standards  for gases  weekly or sooner if
      comparison with check standards indicates a problem.   Reactive compounds
      such as 2-chloroethyl  vinyl ether  and styrene may need to be prepared more
      frequently.  All other standards  must be  replaced after six months.  Both
      gas and liquid  standards must  be  monitored  closely  by comparison  to the
      initial calibration curve and  by comparison to QC check standards.  It may
      be  necessary to replace the  standards  more  frequently if  either check
      exceeds a 20% drift.

      5.6   Prepare   secondary  dilution  standards,  using   stock  standard
solutions, in methanol, as needed, that contain the compounds of interest, either
singly or mixed together.  The secondary dilution standards should be prepared
at concentrations such that  the aqueous calibration standards prepared in Sec.
5.7 will  bracket the working range of the analytical  system.  Secondary dilution
standards should be stored with  minimal  headspace  for  volatiles  and should be
checked frequently for signs  of degradation.or evaporation, especially just prior
to preparing calibration standards  from them.

      5.7   Calibration standards,  at a  minimum of five concentration levels are
prepared in organic-free reagent  water  from  the secondary dilution of the stock
standards.  One of the concentration levels should be at  a concentration near,
but above, the method detection limit.  The remaining concentration levels should
correspond to the  expected range  of  the  concentrations found in real samples or
should define  the working range of the GC.   Standards  (one or  more)  should
contain each  analyte  for detection by this method.   In order to prepare accurate
aqueous standard solutions,  the following precautions must  be observed.

      NOTE: Prepare calibration solutions for use with direct injection analyses
            in water  at the concentrations  required.

            5.7.1  Do  not inject  more  than  20  /xL  of alcoholic  standards  into
      100 ml of water.

            5.7.2  Use   a  25  /uL Hamilton  702N  microsyringe  or  equivalent
      (variations  in  needle  geometry   will  adversely  affect  the  ability  to
      deliver reproducible volumes of methanolic standards into water).

            5.7.3  Rapidly   inject   the   alcoholic   standard  into  the  filled
      volumetric flask.   Remove the  needle as fast as possible after injection.

            5.7.4  Mix aqueous  standards by  inverting the flask three  times.
                                  8021A  - 6                         Revision 1
                                                                September 1994

-------
             5.7.5  Fill  the  sample  syringe  from the standard solution contained
      in the  expanded area  of  the  flask (do not  use any solution contained in
      the neck of the flask).

             5.7.6  Never use pipets  to  dilute or transfer  samples  or aqueous
      standards.

             5.7.7  Aqueous standards are not stable and should be discarded after
      one hour,  unless  properly  sealed  and stored.   The aqueous standards can
      be stored  up to 12 hours,  if held in  sealed vials with zero headspace.

             5.7.8  Optionally calibration using a certified gaseous mixture can
      be accomplished  daily utilizing  commercially  available  gaseous analyte
      mixture  of bromomethane,  chloromethane,  chloroethane,  vinyl  chloride,
      dichlorodifluoromethane  and  trichlorofluoromethane  in  nitrogen.  These
      mixtures  of documented  quality  are  stable  for  as  long as  six months
      without  refrigeration. (VOA-CYL  III,  RESTEK  Corporation, Cat.  #20194 or
      equivalent).

      5.8    Internal  standards  -   Prepare  a   spiking  solution  containing
fluorobenzene  and  2-bromo-l-chloropropane  in methanol, using  the  procedures
described in Sees. 5.5 and 5.6.  It is recommended that the secondary dilution
standard be  prepared  at a concentration of 5 mg/L of  each  internal  standard
compound.   The addition of  10 /jL  of such  a  standard to 5.0  ml  of  sample or
calibration standard would be equivalent to 10 jug/L.

      5.9    Surrogate   standards  -   The   analyst   should   monitor  both  the
performance  of  the  analytical  system and  the effectiveness of the  method in
dealing with each sample matrix  by spiking  each  sample, standard,  and reagent
blank with  two or more surrogate compounds.   A  combination of 1,4-dichlorobutane
and bromochlorobenzene is recommended to encompass the range of the temperature
program used in this  method.  From  stock standard solutions prepared as in Sec.
5.5, add a  volume  to give 750 /ig  of each  surrogate  to 45  ml  of  organic-free
reagent water contained in a 50 ml  volumetric flask,  mix, and dilute to volume
for a concentration of 15 ng/juL.   Add 10 /LtL of this surrogate spiking solution
directly into the 5 ml syringe with every sample and reference standard analyzed.
If the internal standard calibration  procedure is used,  the surrogate compounds
may be added directly to the internal standard spiking  solution (Sec. 5.8).


6.0   SAMPLE COLLECTION, PRESERVATION, AND  HANDLING

      6.1    See  the  introductory material   to this chapter,  Organic  Analytes,
Sec. 4.1.


7.0   PROCEDURE

      7.1    Volatile compounds are  introduced into the gas chromatograph either
by direct injection or  purge-and-trap (Method 5030).  Method  5030  may be used
directly on  ground water samples or low-concentration  contaminated  soils  and
sediments.   For medium-concentration  soils or sediments,  methanolic extraction,
as described in Method 5030,  may  be necessary prior to purge-and-trap analysis.


                                  8021A -  7                        Revision 1
                                                                September 1994

-------
      7.2   Gas chromatography conditions  (Recommended)

            7.2.1  Set   up   the   gas   chromatograph  system   so   that   the
      photoionization  detector  (PID)   is   in   series  with  the  electrolytic
      conductivity detector  (HECD).

            7.2.2  Oven  settings:

            Carrier gas (Helium) Flow rate:     6 mL/min.
            Temperature program
                   Initial  temperature:     10°C,  hold for 8 minutes at
                   Program:                 10°C  to 180°C  at  4°C/miri
                   Final temperature:       180°C,  hold   until    all   expected
                                           compounds  have eluted.

            7.2.3  The carrier gas flow is augmented  with  an additional 24 ml of
      helium flow  before  entering  the photoionization  detector.   This make-up
      gas is necessary to ensure optimal response from both detectors.

            7.2.4  These halogen-specific systems eliminate misidentifications
      due to non-organohalides which are coextracted during the purge step.  A
      Tracer Hall  Model  700-A detector was  used  to gather the single laboratory
      accuracy  and  precision data  presented   in  Table  2.    The  operating
      conditions used to collect these data are:

            Reactor tube:                       Nickel,  1/16  in OD
            Reactor temperature:                810°C
            Reactor base  temperature:           250°C
            Electrolyte:                        100% n-Propyl  alcohol
            Electrolyte flow rate:              0.8 mL/min
            Reaction gas:                       Hydrogen at 40 mL/min
            Carrier gas plus make-up gas:       Helium at 30  mL/min

            7.2.5  A  sample  chromatogram  obtained with this column is presented
      in  Figure  5.  This  column was used to  develop the method performance
      statements in Sec.  9.0.  Estimated retention  times and  MDLs that can be
      achieved under these  conditions are  given in  Table 1.   Other columns or
      element specific detectors  may be used if the requirements of Sec. 8.0 are
      met.

      7.3   Calibration   -   Refer  to   Method   8000  for  proper  calibration
techniques.   Use Table 1  and especially  Table 2 for guidance  on selecting the
lowest point on the calibration curve.

            7.3.1  Calibration must take  place using the same sample introduction
      method that will be used to analyze  actual samples (see Sec. 7.4.1).

            7.3.2  The  procedure  for internal  or external calibration  may be
      used.   Refer  to Method 8000 for a description of each  of these procedures.

      7.4   Gas chromatographic analysis
                                  8021A  - 8                         Revision 1
                                                                September 1994

-------
             7.4.1  Introduce volatile compounds into the gas chromatograph using
      either Method 5030 (purge-and-trap method)  or the direct injection method
      (see  Sec.  7.4.1.1).   If the  internal  standard  calibration technique is
      used,  add  10 /nL  of  internal  standard  to  the  sample  prior  to  purging.

                   7.4.1.1     Direct injection -  In very  limited applications
             (e.g.  aqueous  process  wastes)  direct injection  of the sample into
             the  GC  system with a 10  juL   syringe  may be  appropriate.    The
             detection limit is very high (approximately 10,000 Mg/L), therefore,
             it is only permitted where concentrations  in excess of 10,000 /Ltg/L
             are  expected or for water-soluble compounds that do  not purge.  The
             system must be calibrated by direct injection  (bypassing the purge-
             and-trap device).

             7.4.2  Follow   Sec.  7.6  in  Method 8000  for  instructions on  the
      analysis sequence,  appropriate dilutions,  establishing  daily  retention
      time  windows,  and  identification  criteria.   Include a mid-concentration
      standard after each  group of  10 samples  in the  analysis sequence.

             7.4.3  Table  1  summarizes  the estimated retention  times on the two
      detectors for a number of organic  compounds analyzable using this method.

             7.4.4  Record  the  sample volume  purged or injected and the resulting
      peak  sizes (in area  units or peak heights).

             7.4.5  Calculation of concentration is  covered in Method 8000.

             7.4.6  If analytical interferences are suspected, or  for the purpose
      of confirmation, analysis using a second GC column is recommended.

             7.4.7  If the  response  for  a peak is off-scale,  i.e.,  beyond the
      calibration range of the standards, prepare a  dilution of the sample with
      organic-free reagent water.   The  dilution  must  be performed  on  a second
      aliquot of the sample which has been  properly sealed and stored  prior to
      use.
8.0   QUALITY CONTROL

      8.1   Refer  to  Chapter One for specific quality  control  procedures and
Method 8000 for gas chromatographic procedures.   Quality control to ensure the
proper operation of the purge-and-trap device is  covered in Method 5030.

      8.2   Quality control  required to validate  the  GC  system  operation  is
found in Method 8000.

            8.2.1  The  quality control  reference  sample (Method  8000)  should
      contain  each parameter of  interest  at  a  concentration  of 10 mg/L  in
      methanol.

            8.2.2  Table  2  gives  method  accuracy  and  precision as functions  of
      concentration for the analytes of interest.
                                  8021A  - 9                         Revision 1
                                                                September 1994

-------
      8.3   Calculate  surrogate  standard  recovery  on  all  samples,  blanks, and
spikes.  Determine if recovery is within limits (limits  established by performing
QC procedure outlined in Method 8000).

            8.3.1  If  recovery  is not within limits, the following is required.

                   •     Check to  be sure there are  no errors in calculations,
                        surrogate solutions  and internal standards.  Also check
                        instrument  performance.

                   •     Recalculate the data and/or reanalyze the extract  if any
                        of the  above checks  reveal  a problem.

                   •     Reextract and  reanalyze the  sample if none of the above
                        are  a  problem   or  flag   the  data   as   "estimated
                        concentration".
9.0   METHOD PERFORMANCE

      9.1   Method detection limits for these analytes have been calculated from
data collected by spiking organic-free reagent water  at  0.1  fj.g/1.   These data
are presented in Table 1.

      9.2   This  method  was tested in a  single  laboratory using organic-free
reagent water spiked at 10 jug/L.   Single laboratory precision and accuracy data
for each detector are presented for the method analytes in Table 2.


10.0  REFERENCES

1.    Volatile Organic Compounds^in Hater by Purqe-and-Trap Capillary Column Gas
      Chromatography   with   Photoionization  and   Electrolytic   Conduct ivity
      Detectors  in  Series.  Method 502.2.  Rev.  2.0  (1989)
      Determination  of Organic  Compounds  in
      Monitoring Systems Laboratory, Cincinnati
      1988
                                                        Methods for  the
                                         Drinking Water",  Environmental
                                        ,  OH, EPA/600/4-88/039, December,
      The Determination of Halogenated Chemicals  in Water by the Purge and Trap
      Method,  Method  502.1;  Environmental  Protection  Agency,  Environmental
      Monitoring  and Support Laboratory:  Cincinnati,  Ohio 45268,  September,
      1986.

      Volatile Aromatic and Unsaturated Organic Compounds in Water by Purge and
      Trap Gas Chromatography, Method  503.1;  Environmental  Protection Agency,
                                               Laboratory:
Environmental  Monitoring
September, 1986.
and  Support
Cincinnati,  Ohio,
      Glaser, J.A.; Forest, D.L.; McKee, G.D.; Quave, S.A.; Budde, W.L. "Trace
      Analyses for Wastewaters"; Environ. Sci. Technol.  1981, 15, 1426.
                                  8021A - 10
                                                              Revision  1
                                                          September 1994

-------
5.     Bellar, T.A.;  Lichtenberg,  J.J. The  Determination  of Synthetic Organic
      Compounds In Water by Purge and Sequential Trapping Capillary Column Gas
      Chromatoqraph.y;  U.S.  Environmental  Protection  Agency,  Environmental
      Monitoring and Support Laboratory: Cincinnati, Ohio,  45268.
                                  8021A  -  11                         Revision  1
                                                                 September  1994

-------
                              TABLE  1.

CHROMATOGRAPHIC RETENTION TIMES AND METHOD DETECTION LIMITS (MDL) FOR
  VOLATILE ORGANIC COMPOUNDS ON PHOTOIONIZATION DETECTION (PID) AND
      HALL  ELECTROLYTIC  CONDUCTIVITY  DETECTOR  (HECD) DETECTORS
Analyte
Dichlorodifluoromethane
Chloromethane
Vinyl Chloride
Bromomethane
Chloroethane
Tri chl orof 1 uoromethane
1,1-Dichloroethene
Methylene Chloride
trans-l,2-Dichloroethene
1,1-Dichloroethane
2,2-Dichloropropane
cis-l,2-Di chloroethane
Chloroform
Bromochloromethane
1 ,1,1 -Tri chloroethane
1,1-Dichloropropene
Carbon Tetrachloride
Benzene
1,2-Dichloroethane
Trichloroethene
1,2-Dichloropropane
Bromod ichl oromethane
Dibromomethane
Toluene
1,1, 2 -Tri chloroethane
Tetrachloroethene
1,3-Dichloropropane
Dibromochl oromethane
1,2-Dibromoethane
Chlorobenzene
Ethylbenzene
1,1,1 , 2-Tetrachloroethane
m-Xylene
p-Xylene
o-Xylene
Styrene
Isopropyl benzene
Bromoform
1,1, 2, 2-Tetrachloroethane
1,2,3-Trichloropropane
PID
Ret. Time8
minute
b
-
9.88
-
-
-
16.14
-
19.30
-
-
23.11
-
-
-
25.21
-
26.10
-
27.99
-
-
-
31.95
-
33.88
-
-
-
36.56
36.72
-
36.98
36.98
38.39
38.57
39.58
-
-
-
HECD
Ret. Time
minute
8.47
9.47
9.93
11.95
12.37
13.49
16.18
18.39
19.33
20.99
22.88
23.14
23.64
24.16
24.77
25.24
25.47
-
26.27
28.02
28.66
29.43
29.59
-
33.21
33.90
34.00
34.73
35.34
36.59
-
36.80
-
-
-
-
-
39.75
40.35
40.81
PID
MDL
M9/L


0.02



NDC

0.05


0.02



0.02

0.009

0.02



0.01

0.05



0.003
0.005

0.01
0.01
0.02
0.01
0.05



HECD
MDL
M9/L
0.05
0.03
0.04
1.1
0.1
0.03
0.07
0.02
0.06
0.07
0.05
0.01
0.02
0.01
0.03
0.02
0.01

0.03
0.01
0.006
0.02
2.2

ND
0.04
0.03
0.03
0.8
0.01

0.005





1.6
0.01
0.4
                             8021A  -  12
    Revision 1
September 1994

-------
                                   TABLE 1.
                                  (Continued)
Analyte
   PID
Ret.  Time8
  minute
  HECD
Ret. Time
 minute
PID
MDL
M9/L
HECD
 MDL
M9/L
n-Propylbenzene                   40.87
Bromobenzene                      40.99
1,3,5-Trimethylbenzene            41.41
2-Chlorotoluene                   41.41
4-Chlorotoluene                   41.60
tert-Butylbenzene                 42.92
1,2,4-Trimethylbenzene            42.71
sec-Butyl benzene                  43.31
p-Isopropyltoluene                43.81
1,3-Dichlorobenzene               44.08
1,4-Dichlorobenzene               44.43
n-Butylbenzene                    45.20
1,2-Dichlorobenzene               45.71
l,2-Dibromo-3-Chloropropane
1,2,4-Trichlorobenzene            51.43
Hexachlorobutadiene               51.92
Naphthalene                       52.38
1,2,3-Trichlorobenzene            53.34

Internal Standards
  Fluorobenzene                   26.84
  2-Bromo-l-chloropropane
                 41.03

                 41.45
                 41.63
                 44.11
                 44.47

                 45.74
                 48.57
                 51.46
                 51.96

                 53.37
                 33.08
             0.004
             0.006
             0.004
             ND
             0.02
             0.06
             0.05
             0.02
             0.01
             0.02
             0.007
             0.02
             0.05

             0.02
             0.06
             0.06
             ND
          0.03

          0.01
          0.01
          0.02
          0.01

          0.02
          3.0
          0.03
          0.02

          0.03
    Retention times determined  on  60 m  x  0.75 mm  ID
    Program:  Hold at 10°C for 8 minutes, then  program
    hold until  all  expected compounds have  eluted.

    Dash (-)  indicates  detector does not respond.

    ND = Not  determined.
                     VOCOL capillary  column.
                     at  4°C/min  to  180°C,  and
                                  8021A - 13
                                   Revision  1
                               September  1994

-------
                   TABLE 2.
SINGLE LABORATORY ACCURACY AND PRECISION DATA
   FOR VOLATILE ORGANIC COMPOUNDS IN WATERd
Photoionization
Detector
Analyte
Benzene
Bromobenzene
Bromochloromethane
Bromodichloromethane
Bromoform
Bromomethane
n-Butyl benzene
sec-Butyl benzene
tert-Butyl benzene
Carbon tetrachloride
Chlorobenzene
Chloroethane
Chloroform
Chloromethane
2-Chlorotoluene
4-Chlorotoluene
1 ,2-Dibromo-3-chloropropane
Dibromochloromethane
1,2-Dibromoethane
Dibromomethane
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
Dichlorodifluoromethane
1, 1-Dichloroethane
1,2-Dichloroethane
1 ,1-Dichloroethene
cis-1,2 Dichloroethene
trans-l,2-Dichloroethene
1 ,2-Dichloropropane
1 ,3-Dichloropropane
2,2-Dichloropropane
1 , 1-Dichloropropene
Ethyl benzene
Hexachlorobutadiene
I sopropyl benzene
p-Isopropyltoluene
Recovery,8
%
99
99
-
-
-
-
100
97
98
-
100
-
-
-
NDC
101
-
-
-
-
102
104
103
-
-
-
100
ND
93
-
-
-
103
101
99
98
98
Standard
Deviation
of Recovery
1.2
1.7
-
-
-
-
4.4
2.6
2.3
-
1.0
-
-
-
ND
1.0
-
-
-
-
2.1
1.7
2.2
-
-
-
2.4
ND
3.7
-
-
-
3.6
1.4
9.5
0.9
2.4
Hall Electrolytic
Conductivity Detector
Standard
Recovery,8 Deviation
% of Recovery
_b
97
96
97
106
97
-
-
-
92
103
96
98
96
97
97
86
102
97
109
100
106
98
89
100
100
103
105
99
103
100
105
103
-
98
-
-
.
2.7
3.0
2.9
5.5
3.7
-
-
-
3.3
3.7
3.8
2.5
8.9
2.6
3.1
9.9
3.3
2.7
7.4
1.5
4.3
2.3
5.9
5.7
3.8
2.9
3.5
3.7
3.8
3.4
3.6
3.4
-
8.3
-
-
                  8021A - 14
    Revision 1
September 1994

-------
                                         TABLE 2.
                                        (Continued)
Analyte
Photoionization
Detector
Standard
Recovery,8 Deviation
% of Recovery
Hall Electrolytic
Conductivity Detector
Standard
Recovery,8 Deviation
% of Recovery
Methylene chloride
Naphthalene
n-Propylbenzene
Styrene
1,1, 1,2-Tetrachloroethane
1,1,2,2-Tetrachloroethane
Tetrachloroethene
Toluene
1,2,3-Trichlorobenzene
1,2,4-Trichlorobenzene
1,1, 1-Trichloroethane
1,1,2-Trichloroethane
Trichloroethene
Trichlorofluoromethane
1,2,3-Trichloropropane
1,2, 4 -Tri methyl benzene
1,3, 5 -Tri methyl benzene
Vinyl chloride
o-Xylene
m-Xylene
p-Xylene

102
103
104
-
-
101
99
106
104
-
-
100
-
-
99
101
109
99
100
99

6.3
2.0
1.4
-
-
1.8
0.8
1.9
2.2
-
-
0.78
-
-
1.2
1.4
5.4
0.8
1.4
0.9
97
-
-
-
99
99
97
-
98
102
104
109
96
96
99
-
-
95
-
-
~
2.8
-
-
-
2.3
6.8
2.4
-
3.1
2.1
3.4
6.2
3.5
3.4
2.3
-

5.6

-
~
    Recoveries  and standard  deviations  were determined  from  seven samples  and  spiked at
    10  /xg/L of each analyte. Recoveries were  determined by internal  standard method. Internal
    standards were:  Fluorobenzene  for PID,  2-Bromo-l-chloropropane for HECD.
b   Detector  does  not  respond.

0   ND  =  Not  determined.
    This  method  was  tested
    reference  5).
in a  single laboratory  using water  spiked at  10
                (see
                                        8021A - 15
    Revision 1
September 1994

-------
                      TABLE 3.
DETERMINATION OF ESTIMATED QUANTITATION LIMITS (EQL)
                FOR VARIOUS  MATRICES3
  Matrix                              Factor
  Ground water                             10
  Low-concentration soil                   10
  Water miscible liquid waste             500
  High-concentration soil and sludge     1250
  Non-water miscible waste               1250
  EQL = [Method detection limit (see Table 1)] X [Factor found in
  this table].  For non-aqueous samples,  the  factor  is  on a wet-
  weight basis.  Sample EQLs are highly matrix-dependent.  The EQLs
  listed herein are provided  for  guidance and may not  always be
  achievable.
                     8021A - 16                        Revision 1
                                                   September 1994

-------
                            FIGURE  1.
                         PURGING  DEVICE
OPTIONM.
POAMTTM*
                        OUT i<4 M 0.0
•— 14 MM 00
                        INLET 1M IN 00
Exrr IK IN. o.o
10 MM GLASS FRTT
MEDIUM
           SAMPLf INLET


           J4VAY SYAINGC VALVC

           17 CM 20 GAUGE SVKNGC NEEDLE


            MM 0 0 RU8KR SCFTUM

           INLET 1M IN 0 0
                                                      1'ie IN 00
                                                  ^ STAINLESS STEEL
                                                     '3X
                                                     MOLECULAR SIEVE
                                                           GAS
                                                            GAS
                                                       Fl.CV< CONTHOt
                          8021A  -  17
                                                 Revision  1
                                            September 1994

-------
                           FIGURE 2.
TRAP PACKINGS AND CONSTRUCTION TO INCLUDE  DESORB CAPABILITY
       PACKING DETAIL


             i UUO>38 "OX
             CM S4UO G£L
                                   CONSTRUCTOR OFTAJL
                    3C
=T
             • CM JX Ov-t
             5 MM il>S8 WOOL
                                                n*iNG a CM
                                                a'06 IN. iD
                                                0 '» it*. 00.
                         8021A  -  18
                                                    Revision  1
                                                September  1994

-------
                            FIGURE 3.
               PURGE-AND-TRAP  SYSTEM -  PURGE MODE
CAPWERGAS
FLOW CONTROL
— uowo INJECTION PO«TS
        COLUMN OVEN
                                                              COCUMN
PXWGCGAS
PLOW CONTROL
SIEVE RLTER
                                                           COLUMN
                               OPTIONAL 4^O*TT COLUMN
                               SELKTDON VALVE
                                ».    /- TRAP INLET
                                       TRAP
                                       22 *C
                                                NOTE.
                                                ALL UNtS 8€TWEEN
                                                ANO QC SHOULD K H€AfEO
                                                TO«3*C.
                             8021A -  19
                                Revision 1
                            September 1994

-------
                            FIGURE  4.
        SCHEMATIC  OF  PURGE-ANO-TRAP DEVICE - DESORB MODE
CARRIER GAS
FLOW CONTROL
PRESSURE
REGULATOR
PURGE GAS
FLOW CONTROL
13X MOLECULAR
SIEVE FILTER
r- UOWO INJECTION PORTS

}      r— COLUMN OVEN

                                                 CONFIRMATORY COLUMN
                                                TO DETECTOR
                                                 ANALYTICAL COLUMN
                               OPTIONAL 4PORT COLUMN
                               SELECTION VALVE
                                        TRAP INLET
                                       TRAP
                                       200*C
if PURGING
 1 OCVCC
                                               won.
                                               ALL UNES BETWEEN TRAP
                                               AMD GC SHOULD BE HEATED
                            8021A  -  20
                                Revision  1
                            September 1994

-------
                                   FIGURE 5.
                     GAS CHROMATOGRAM OF VOLATILE ORGANICS
COLUMN I 60 METER M 0. 73 MM  I.D.  VOCOL CAPILLARY

     AM) TWA* VOC'l WITH MM.L t *IO IN SCRIES
;= n   sa
s  t
                                      2*28
Sf^S
8  S
                                                                        ^v wtinnHt n
       1
                            JU
                                                     JUUUL PID
                                                                                    KSCO
                                  8021A  -  21
                                             Revision 1
                                         September 1994

-------
                                      METHOD  8021A
       HALOGENATED VOLATILES  BY GAS CHROMATOGRAPHY USING PHOTOIONIZATION
               AND ELECTROLYTIC  CONDUCTIVITY DETECTORS  IN  SERIES:
                              CAPILLARY  COLUMN TECHNIQUE
 (     Start     J
      7.2 Set
  chromatographic
     conditions.
    7.3 Refer to
  Method 8000 for
calibration techniques.
  7.4.1  Introduce
sample into GC using
  direct injection or
  purge-snd-trap.
    7.4.4 Record
   sample volume
 introduced into GC
   and peak sizes.
   7.4.5 Refer
to Method 8000 for
   calculations.
      7.4.6 Are
      analytical
    interferences
     suspected?
                                      7.4.7 Is peak
                                      response off
                                         scale?
Reanalyze sample
using second GC
    column.
                             Dilute and reanalyze
                              second aliquot of
                                   sample.
                                        8021A  -  22
                                               Revision 1
                                          September  1994

-------
00
o

-------
                                 METHOD 8030A

               ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY


1.0   SCOPE AND APPLICATION

      1.1   Method 8030 is used to determine the concentration of the following
volatile organic compounds:
            Compound Name                                   CAS No.8


            Acrolein (Propenal)                             107-02-8
            Acrylonitrile                                   107-13-1


      a  Chemical  Abstract Services Registry Number.

      1.2   Table 1 lists  chromatographic conditions and method detection limits
for acrolein and acrylonitrile  in organic-free reagent water.  Table 2 lists the
estimated quantitation limit (EQL) for other matrices.


2.0   SUMMARY OF METHOD

      2.1   Method 8030 provides gas chromatographic conditions for the detection
of the target analytes.  Samples can be analyzed using direct injection or purge-
and-trap  (Method  5030).   Tenax should  be  used as the trap  packing material.
Ground water samples must be analyzed  using  Method 5030.  A temperature program
is used in the gas chromatograph to separate the organic compounds.   Detection
is achieved by a flame ionization detector  (FID).

      2.2   The method provides an optional  gas chromatographic column that may
be helpful in resolving the analytes from co-eluting non-target compounds and for
analyte confirmation.


3.0   INTERFERENCES

      3.1   Refer to Methods 5030 and 8000.

      3.2   Samples  can   be  contaminated  by  diffusion  of  volatile  organics
(particularly chlorofluorocarbons  and methylene chloride) through  the  sample
container  septum  during  shipment  and  storage.   A  trip blank  prepared  from
organic-free reagent water and carried through sampling and subsequent storage
and handling can serve as a check on such contamination.
                                   8030A -  1                         Revision 1
                                                                     July 1992

-------
4.0   APPARATUS AND MATERIALS

      4.1   Gas chromatograph

            4,1.1      Gas Chromatograph - Analytical  system complete with gas
      chromatograph suitable for on-column injections  or purge-and-trap sample
      introduction and  all  required accessories,  including  detectors,  column
      supplies, recorder,  gases,  and syringes.  A data  system  for measuring peak
      height and/or peak area is recommended.

            4.1.2      Columns

                 4.1.2.1      Column 1  -  10 ft  x  2 mm ID stainless  steel  or
            glass packed with Porapak-QS (80/100 mesh) or equivalent.

                 4.1.2.2      Column 2  - 6  ft  x  0.1 in.  ID stainless steel  or
            glass packed with Chromosorb 101 (60/80 mesh) or equivalent.

            4.1.3      Detector - Flame ionization (FID).

      4.2   Sample  introduction apparatus  -  Refer   to  Method  5030 for  the
appropriate equipment for sample introduction purposes.

      4.3   Syringes - A  5  mL  Luer-lok glass hypodermic  and  a 5 ml, gas-tight
with shutoff valve.

      4.4   Volumetric  flasks,  Class  A - Appropriate  sizes  with ground glass
stoppers.

      4.5   Microsyringes   -  10  and   25  pi  with   a  0.006  in.  ID  needle
(Hamilton 702N, or equivalent) and a 100 /iL.

      4.6   Analytical balance - 0.0001 g.


5.0   REAGENTS

      5.1   Reagent grade chemicals'shall  be used in all tests.  Unless otherwise
indicated, it is intended that all  reagents  shall  conform to  the  specifications
of the Committee on Analytical Reagents of the American Chemical  Society, where
such specifications are available. Other grades may be  used, provided it is first
ascertained that  the  reagent is of sufficiently high  purity to permit  its use
without  lessening the accuracy of the  determination.

      5.2   Organic-free  reagent water:  All references to  water in this method
refer to organic-free reagent water, as defined  in Chapter One.

      5,3   Hydrochloric  acid,  HC1  - 1:1  (v/v).

      5.4   Sodium  hydroxide, NaOH  -   ION  solution.    Dissolve  40 g  NaOH in
organic-free reagent water  and dilute  to  100 mL.

      5.5   Stock standards  - Stock solutions may be prepared from pure standard
materials  or purchased  as   certified  solutions.   Prepare  stock standards in

                                   8030A - 2                         Revision  1
                                                                     July  1992

-------
organic-free  reagent  water  using  assayed  liquids.    Because  acrolein  and
acrylonitrile are lachrymators, primary dilutions of these compounds should be
prepared in a hood.

            5.5.1      Place about 9.8  ml  of  organic-free reagent water in a 10
      ml tared ground-glass stoppered volumetric  flask.   For acrolein standards
      the water must be adjusted to pH  4-5 using  hydrochloric acid (1:1 v/v) or
      sodium hydroxide  (ION),, if necessary.   Weigh the  flask to  the nearest
      0.0001 g.

            5.5.2      Using a 100 /iL syringe, immediately add two or more drops
      of assayed reference material  to  the flask,  then reweigh.  The liquid must
      fall directly into the water without contacting the neck of the flask.

            5.5.3      Reweigh, dilute to volume,  stopper,   and  then mix  by
      inverting  the flask  several times.    Calculate  the  concentration  in
      milligrams per liter  (mg/L)  from the net gain in weight.   When compound
      purity is  assayed  to  be 96% or   greater, the  weight may  be  used without
      correction  to  calculate  the   concentration  of  the  stock  standard.
      Commercially prepared stock standards may be used at any concentration, if
      they are certified by the manufacturer or by an independent source.

            5.5.4      Transfer the stock standard solution  into a bottle with
      a Teflon  lined screw-cap.   Store,  with minimal   headspace, at  4°C  and
      protect from light.

            5.5.5      Prepare fresh standards daily.

      5.6   Secondary dilution standards - Prepare secondary dilution standards
as needed,  in organic-free reagent water, from the stock standard solutions.  The
secondary  dilution  standards must  contain  the compounds of  interest,  either
singly or mixed together.  The secondary dilution standards  should be prepared
at concentrations such that the aqueous  calibration standards prepared in Section
5.7 will bracket the working range of the analytical  system.  Secondary dilution
standards  should be  stored  with  minimal  headspace,  and  should be  checked
frequently for  signs  of degradation or evaporation,  especially just  prior to
preparing calibration standards from them.

      5.7   Cal ibration standards - Prepare cal ibration standards in organic-free
reagent water  from  the secondary  dilution   standards  at a  minimum of five
concentrations.  One of the concentrations  should  be at a concentration near, but
above,  the method  detection  limit.    The   remaining   concentrations  should
correspond to the  expected  range of concentrations  found in  real  samples,  or
should define the working range of the GC.   Each standard should  contain each
analyte for detection  by this method.   In order to  prepare  accurate aqueous
standard solutions, the following precautions must be observed.

            5.7.1      Use a 25 pi  Hamilton  702N microsyringe,  or equivalent,
      (variations  in  needle  geometry   will  adversely affect  the ability  to
      deliver reproducible volumes of   standards into water).

            5.7.2      Never use pipets to dilute  or transfer samples or aqueous
      standards.


                                   8030A - 3                         Revision 1
                                                                     July 1992

-------
            5.7.3      Standards must be prepared daily.

      5.8   Internal standards (if internal standard calibration is used) - To
use this approach, the analyst must select one or more internal  standards that
are similar in analytical behavior to  the  compounds  of interest.   The analyst
must further demonstrate that the measurement  of the internal  standard is not
affected by method or  matrix  interferences.   Because of these limitations, no
internal standard can be suggested that is applicable to all samples.

            5.8.1      Prepare  calibration  standards  at  a minimum of  five
      concentrations for each parameter of interest,  as described  in Section
      5.7.

            5.8.2      Prepare a spiking solution containing each of the internal
      standards,  using the  procedures described in Sections  5.5  and 5.6.  It is
      recommended  that  the   secondary  dilution  standard   be   prepared  at  a
      concentration of 15 mg/L of each internal standard compound.  The addition
      of 10 ML of this  standard  to  5.0 ml of  sample  or calibration standard
      would be equivalent to 30
            5.8.3      Analyze each calibration standard  according  to Section
      7.0, adding 10 ML  of  internal  standard  spiking solution directly to the
      syringe.

      5.9   Surrogate standards - The analyst should monitor both the performance
of the analytical  system and the effectiveness  of the method in dealing with each
sample matrix by spiking each sample, standard, and organic-free reagent water
blank with one or two surrogate compounds  (e.g. compounds  similar in analytical
behavior to the analytes  of  interest but which are  not expected to be present in
the sample) recommended to encompass the range of the temperature program used
in this method.  From stock standard solutions prepared as in Section 5.5, add
a volume to give 750 /ug of each surrogate to 45 mL  of  organic-free reagent water
contained  in   a  50  mL  volumetric  flask,  mix,  and  dilute   to  volume for  a
concentration  of 15 ng//iL.   Add  10 nl  of  this surrogate  spiking  solution
directly into  the 5 mL syringe with every sample and reference  standard analyzed.
If the internal standard  calibration procedure is  used, the surrogate compounds
may be added directly to the internal standard  spiking solution (Section 5.8.2).


6.0   SAMPLE COLLECTION,  PRESERVATION, AND HANDLING

      6.1   See  the  introductory  material  to  this  chapter,  Organic Analytes,
Section 4.1.
7.0   PROCEDURE

      7.1   Volatile compounds are introduced into the gas chromatograph either
by direct injection or heated purge-and-trap (Method  5030).  Method 5030 may be
used directly on ground water samples or low-concentration contaminated soils and
sediments.  For high-concentration soils or sediments, methanolic extraction, as
described in Method 5030, may be necessary prior to purge-and-trap analysis.
                                   8030A -  4                         Revision 1
                                                                     July 1992

-------
      7.2   Gas chromatographic conditions (Recommended)

            7.2.1      Column 1:

             Helium flow rate =                 30 mL/min
             Temperature program:
                       Initial  temperature -    110°C,  hold for 1.5 minutes
                       Program -                 110°C  to  150°C,   heating  as
                                                rapidly as possible
                       Final  temperature =      150°C,  hold for 20 minutes.

            7.2.2      Column 2:

             Helium flow rate =                 40 mL/min
             Temperature program:
                       Initial  temperature -    80°C, hold for 4 minutes
                       Program -                 80°C to 120°C  at 50°C/min
                       Final  temperature »      120°C,  hold for 12 minutes.

      7.3   Calibration - Refer to Method 8000 for proper calibration techniques.
Use Table 1  and especially Table 2 for guidance on selecting the  lowest point on
the calibration curve.

            7.3.1      Calibration  must  take  place  using  the  same  sample
      introduction method  that will'  be used  to  analyze  actual  samples  (see
      Section 7.4.1).

            7.3.2      The  procedure for internal or external calibration may be
      used.   Refer to Method 8000 for a description of each  of these procedures.

      7.4   Gas chromatographic analysis

            7.4.1      Introduce  volatile compounds into the gas chromatograph
      using either Method  5030 (heated purge-and-trap method using Tenax as the
      trap packing material) or  the direct  injection method.   If the internal
      standard calibration  technique is used, add 10 ni  of the internal standard
      to the sample prior to purging.

                 7.4.1.1      Direct  injection  -  In  very limited applications
            (e.g. aqueous process wastes), direct injection of the sample into
            the  GC system  with   a  10  pi syringe  may  be  appropriate.    The
            detection limit is very high (approximately  10,000 M9A)> therefore,
            it is only permitted  when concentrations in  excess  of 10,000  /^g/L
            are expected or for water-soluble compounds that do  not purge.  The
            system must be  calibrated  by direct injection  (bypassing the purge-
            and-trap device).

            7.4.2      Follow Method  8000  for  instructions  on  the  analysis
      sequence,  appropriate  dilutions,  establishing  daily  retention  time
      windows,  and  identification  criteria.    Include   a  mid-concentration
      standard after each group of  10 samples  in the analysis sequence.

            7.4.3      Table 1  summarizes the estimated  retention  times and
      detection limits for a number of organic compounds analyzable using this

                                  8030A - 5                        Revision  1
                                                                     July 1992

-------
      method.  Figure 1 illustrates the chromatographic separation  of acrolein
      and of aeryfonitrile using Column 1.

            7.4.4      Record  the  sample volume  purged or  injected and  the
      resulting peak sizes (in area units or peak heights).

            7.4.5      Calculation  of concentration  is  covered  in Method 8000.

            7.4.6      If analytical  interferences  are suspected,  or for  the
      purpose  of  confirmation,  analysis   using  the   second   GC   column   is
      recommended.

            7.4.7      If the  response  for  a  peak is  off-scale,  prepare  a
      dilution of the  sample with organic-free reagent water.  The dilution must
      be performed on  a  second  aliquot of the  sample  which  has  been properly
      sealed and stored prior to use.


8.0  QUALITY CONTROL

      8.1   Refer to  Chapter  One  for specific  quality  control  procedures  and
Method 8000 for gas chromatographic procedures.   Quality control  to ensure  the
proper operation of the purge-and-trap device is covered in  Method  5030.

      8.2   Procedures to check the GC system operation are found in Method 8000,
Section 8.6.

            8.2.1      The quality control check  sample concentrate (Method 8000,
      Section 8.6) should contain each parameter of  interest at a concentration
      of 25 mg/L in water.

            8.2.2      Table   3  indicates the  calibration  and QC  acceptance
      criteria for this method.  Table  4 gives  single  laboratory  accuracy  and
      precision  for  the  analytes  of interest.    The contents  of  both  Tables
      should be used to evaluate a  laboratory's  ability  to perform and generate
      acceptable data by this method.

      8.3   Calculate surrogate standard recovery on all  samples,  blanks,  and
spikes.  Determine if recovery is within limits (limits established by performing
QC procedure outlined in Method 8000, Section 8.10).

            8.3.1      If recovery  is  not  within  limits,   the  following  is
      required.

            •    Check to be  sure  that  there  are no  errors  in  calculations,
                 surrogate  solutions  and  internal  standards.   Also,  check
                 instrument performance.

            •    Recalculate the data and/or reanalyze the extract if  any of the
                 above checks reveal  a problem.

            •    Reextract and reanalyze the sample if none of the above are a
                 problem or flag the data as "estimated concentration".


                                   8030A - 6                         Revision 1
                                                                     July 1992

-------
9.0   METHOD PERFORMANCE

      9.1   In  a  single  laboratory,   the  average  recoveries  and  standard
deviations presented in Table 4 were obtained using Method 5030.  Seven replicate
samples were analyzed at each spike concentration.

      9.2   The accuracy and  precision obtained will be determined by the sample
matrix, sample introduction technique, and by the calibration procedure used.


10.0  REFERENCES

1.    Bellar, T.A.  and J.J.  Lichtenberg,  J.  Amer.  Water Works Assoc., 66(12).
      pp. 739-744,  1974.

2.    Bellar, T.A.  and J.J. Lichtenberg, "Semi-Automated Headspace Analysis of
      Drinking Waters  and Industrial  Waters  for  Purgeable  Volatile Organic
      Compounds," in Van Hall, ed., Measurement of Organic Pollutants in Water
      and Wastewater, ASTM STP 686, pp. 108-129, 1979.

3.    Development and Application of  Test  Procedures  for Specific Organic Toxic
      Substances  in Wastewaters,  Category  11:  Purgeables  and Category  12:
      Acrolein,  Acrylonitrile,  and  Dichlorodifluoromethane,  Report for  EPA
      Contract 68-03-2635  (in preparation).

4.    Going,  J.,  et  al.,  Environmental  Monitoring Near  Industrial Sites  -
      Acrylonitrile, Office of Toxic Substances, U.S. EPA, Washington, DC, EPA
      560/6-79-003, 1979.

5.    U.S. EPA 40 CFR Part  136, "Guidelines Establishing Test  Procedures for the
      Analysis of Pollutants Under the Clean Water Act; Final Rule and Interim
      Final Rule and Proposed Rule," October 26, 1984.

6.    Kerns, E.H.,  et al. "Determination of Acrolein  and Acrylonitrile in Water
      by Heated Purge and Trap Technique,"  U.S. Environmental  Protection Agency,
      Environmental Monitoring and Support Laboratory, Cincinnati, Ohio 45268,
      1980.

7.    "Evaluation of Method 603,"  Final Report for EPA Contract  68-03-1760 (in
      preparation).
                                   8030A - 7                        Revision  1
                                                                     July  1992

-------
                             TABLE 1.
      CHROMATOGRAPHIC CONDITIONS AND METHOD DETECTION LIMITS
                        Retention time (min)             Method detection
Compound                Col. 1        Col.  2              limit8  (M9/L)
Acrolein
Acrylonitrile
10.6
12.7
8.2
9.8
0.7
0.5
8  Based on using purge-and-trap, Method 5030.
                             TABLE 2.
              DETERMINATION  OF  ESTIMATED  QUANTITATION
                LIMITS (EQLs)  FOR VARIOUS MATRICES"
         Matrix                              Factor6
         Ground water                             10
         Low-concentration soil                   10
         Water miscible liquid waste             500
         High-concentration soil and sludge     1250
         Non-water miscible waste               1250
         Sample EQLs are highly matrix dependent.  The EQLs listed herein
         are provided for guidance and may not always be achievable.

         EQL *  [Method  detection  limit (Table 1)]  X [Factor (Table 2)].
         For non-aqueous samples, the  factor is on a wet-weight basis.
                             8030A -  8                        Revision 1
                                                               July 1992

-------
                                   TABLE 3.
                    CALIBRATION AND QC  ACCEPTANCE CRITERIA8
Analyte
Acrolein
Acrylonitrile
Range
for Q
(M9/L)
45.9 - 54.1
41.2 - 58.8
Limit
for S
(M9/L)
4.6
9.9
Range
for x
(Atg/L)
42.9 - 60.1
33.1 - 69.9
Range
P> Ps
(%)
88-118
71-135
Q     =     Concentration measured in QC check sample, in M9/L.
S     =     Standard deviation of four recovery measurements, in fj.g/1.
R     =     Average recovery for four recovery measurements, in /ig/L.
P, Ps =     Percent recovery measured.

"     Criteria  from  40  CFR  Part  136  for  Method  603  and were  calculated
      assuming a QC check sample concentration of 50 M9/L-
                                   TABLE 4.
                   SINGLE LABORATORY ACCURACY AND PRECISION


Parameter
Acrolein





Acrylonitrile





AW
POTW
Spike
cone.
UgA)
5.0
50.0
5.0
50.0
5.0
100.0
5.0
50.0
20.0
100.0
10.0
100.0
ASTM Type
Average
recovery
(M9/L)
5.2
51.4
4.0
44.4
0.1
9.3
4.2
51.4
20.1
101.3
9.1
104.0
II water.
Standard
deviation
(M9/L)
0.2
0.7
0.2
0.8
0.1
1.1
0.2
1.5
0.8
1.5
0.8
3.2

Prechlori nation secondary effluent
Average
percent
recovery
104
103
80
89
2
9
84
103
100
101
91
104

from a muni

Sample
matrix
AW
AW
POTW
POTW
IW
IW
AW
AW
POTW
POTW
IW
IW

cipal sewage
IW
treatment plant.
Industrial  wastewater containing  an  unidentified  acrolein
reactant.
                                   8030A - 9
                                                  Revision 1
                                                   July 1992

-------
                   Figure 1
Gas Chromatogram of Acrolein and Acrylonitrile
   Column: Po< Ap«fc OS
           HO*C 
-------
                        METHOD  8030A
 ACROLEIN  AND  ACRYLONITRILE BY  GAS  CHROMATOGRAPHY
      Start
   7  1 Introduce
compounds into  ga s
 chromatograph  by
direct injection or
  purge-and-trap
   (Method 5030)
    7 2 Set gas
   chroma tograpn
    cond 111on
   1  3 Calibrate
 (refer to Method
      8000]
  7  4  1 In t r oduce
volatile compounds
     into gas
 chromatograph by
 purg«-and-trap or
 direct injection
7 4 2  Pol low Method
 8000  for analysis
  sequence,  etc
7  4  4 Record volume
purged or injected
  and peak sizes
  ?  4  5 Calculate
  concent ration
 (refer to Method
      8000)
                   Yesl
^
J
7 4 & Analyze using
second CC col umn

7 4 7 Oil
jle second
aliquot of sample -1
                            8030A  - 11
                                          Revision 1
                                           July  1992

-------
00
o

-------
                                  METHOD  8031

                      ACRYLONITRILE  BY  GAS  CHROMATOGRAPHY


1.0   SCOPE AND APPLICATION

      1.1   Method 8031 is used  to determine the concentration of acrylonitrile
in water.  This method may also  be applicable to other matrices.  The following
compound can be determined by this method:
      Compound Name                                         CAS No.'
      Acrylonitrile                                         107-13-1
      8   Chemical  Abstract  Services  Registry  Number.

      1.2   The estimated quantitation limit  of Method 8031 for determining the
concentration of acrylonitrile in water is approximately 10 jug/L.

      1.3   This method  is restricted to use  by or under the  supervision of
analysts  experienced  in  the use  of gas  chromatographs  and  skilled  in  the
interpretation of gas  chromatograms.   Each analyst must demonstrate the ability
to generate acceptable results with this method.


2.0   SUMMARY OF METHOD

      2.1   A measured sample volume is micro-extracted with methyl tert-butyl
ether.   The extract  is  separated  by gas chromatography and measured  with  a
Nitrogen/Phosphorus detector.


3.0   INTERFERENCES

      3.1   Method  interferences  may be  caused  by contaminants  in  solvents,
reagents, glassware, and  other sample processing hardware that leads to discrete
artifacts and/or elevated baselines in gas chromatograms.  All  of these materials
must be  routinely demonstrated to be free from interferences  under the conditions
of the analysis by running laboratory reagent blanks.

      3.2   Samples can be  contaminated by diffusion of volatile organics around
the  septum  seal  into  the  sample during  handling and  storage.   A  field blank
should  be  prepared from  organic-free reagent water  and carried  through  the
sampling and sample handling  protocol to serve as  a  check  on such contamination.

      3.3   Contamination by  carryover can occur whenever  high-concentration and
low-concentration  samples  are sequentially analyzed. To  reduce  carryover,  the


                                   8031 - 1                          Revision 0
                                                                September 1994

-------
sample syringe  must  be rinsed out  between  samples with  solvent.  Whenever an
unusually concentrated  sample is  encountered,  it  should  be followed  by the
analysis of solvent to check for cross contamination.


4.0   APPARATUS AND MATERIALS

      4.1   Gas chromatograph system

            4.1.1 Gas  chromatograph,  analytical  system  complete  with  gas
      chromatograph  suitable  for   on-column   injections   and   all   required
      accessories, including detector,  analytical  columns,  recorder, gases, and
      syringes.  A data system for measuring peak heights and/or peak areas is
      recommended.

            4.1.2 Column:   Porapak Q  -  6 ft.,  80/10  Mesh, glass  column,  or
      equivalent.

            4.1.3 Nitrogen/Phosphorus detector.

      4.2   Materials

            4.2.1 Grab sample bottles - 40 ml VOA bottles.

            4.2.2 Mixing bottles - 90 mL bottle with a Teflon lined cap.

            4.2.3 Syringes - 10 y.1 and 50 ^L.

            4.2.4 Volumetric flask (Class A) - 100 ml.

            4.2.5 Graduated cylinder - 50 ml.

            4.2.6 Pipet (Class A) - 5, 15, and 50 ml.

            4.2.7 Vials - 10 ml.

      4.3   Preparation

            4.3.1 Prepare all materials to be used as described in Chapter 4 for
      volatile organics.


5.0   REAGENTS

      5.1   Reagent grade chemicals shall be used in all  tests. Unless otherwise
indicated, it is intended that all  reagents  shall  conform to  the  specifications
of the Committee on Analytical  Reagents of the American Chemical  Society, where
such specifications are available. Other grades may be used, provided it is first
ascertained that  the  reagent  is  of sufficiently  high purity to permit its use
without lessening the accuracy of the determination.
                                   8031 - 2                         Revision 0
                                                                September 1994

-------
5.2   General

      5.2.1 Methanol, CH3OH -  Pesticide quality, or equivalent.

      5.2.2 Organic-free reagent water.  All references to water in this
method refer to organic-free reagent water, as defined in Chapter One.

      5.2.3 Methyl tert-butyl  ether,  CH3Ot-C4H9  -  Pesticide  quality,  or
equivalent.

      5.2.4 Acrylonitrile, H2C:CHCN,  98%.

5.3   Stock standard solution

      5.3.1 Stock standard solutions  -  Can  be prepared from pure standard
materials  or  can be  purchased as certified  solutions.    Prepare  stock
standards in organic-free reagent water using assayed liquids.

      5.3.2 The stock standard solution may  be  prepared  by volume  or by
weight.   Stock  solutions  must  be replaced after one year,  or sooner if
comparison with the check standards indicates a problem.

      CAUTION:    Acrylonitrile is toxic.   Standard preparation should be
                  performed in a laboratory fume hood.

            5.3.2.1     To prepare the  stock  standard solution by volume:
      inject 10 ul of acrylonitrile (98%)  into a 100 ml volumetric flask
      with a syringe.  Make up to volume with methanol.

            5.3.2.2     To prepare the  stock  standard solution by weight:
      Place  about 9.8 ml  of  organic-free reagent water  into a  10  ml
      volumetric  flask before  weighing  the flask and  stopper. Weigh the
      flask and record the weight to  the nearest  0.0001 g.  Add two drops
      of pure acrylonitrile, using  a  50 ^L  syringe, to  the  flask.   The
      liquid must  fall  directly  into the  water, without  contacting the
      inside wall  of the  flask.   Stopper the  flask and  then reweigh.
      Dilute to  volume  with organic-free  reagent water.   Calculate the
      concentration from the net gain in weight.

5.4   Working standard solutions

      5.4.1 Prepare a minimum of 5 working standard solutions that cover
the range  of analyte concentrations -expected  in the samples.   Working
standards of 20,  40,  60, 80, and 100 j-tg/L may be  prepared by injecting 10,
20, 30,  40,  and 50 jul  of the  stock standard solution prepared  in Sec.
5.3.2.1 into 5 separate 90 ml mixing  bottles  containing 40 ml of organic-
free reagent water.

      5.4.2 Inject  15  ml  of  methyl  tert-butyl  ether  into  each  mixing
bottle,  shake vigorously,  and  let  stand 5  minutes,  or  until  layers have
separated.
                             8031 - 3                         Revision 0
                                                          September 1994

-------
            5.4.3 Remove 5 ml of top layer by  pipet,  and place  in a 10 ml vial.

            5.4.4 Keep all standard solutions below 4°C until  used.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See the  introductory material to this  chapter,  Organic  Analytes,
Sec. 4.1.


7.0   PROCEDURE

      7.1   Sample Extraction

            7.1.1 Pour 40 mL of the  sample into a 90 mL mixing  bottle.  Pipet 15
      ml of Methyl, tert-butyl  ether into  the  mixing bottle.   Shake vigorously
      for about 2 min. and let stand for about 5 min.   Remove about 5 mL of the
      top layer and store in a 10 mL vial.

      7.2   Chromatographic Conditions  (Recommended)
      Carrier Gas (He) flow rate:
      Column Temperature:
      Injection port temperature:
      Detector temperature:
      Detector Current (DC):
      Gases:

      7.3   Calibration of GC
35 mL/min.
180° C,  Isothermal
250° C
250° C
18 volts
Hydrogen,  3 mL/min;  Air, 290 mL/min.
            7.3.1 On  a  daily basis,  inject  3 /xL  of methyl  tert-butyl  ether
      directly into  the  GC to flush  the  system.   Also purge  the  system with
      methyl tert-butyl  ether injections  between injections  of  standards and
      samples.

            7.3.2 Inject 3  ^L of  a  sample blank (organic-free  reagent  water
      carried through the  sample  storage  procedures  and  extracted  with methyl
      tert-butyl  ether).

            7.3.3 Inject 3 yuL of  at  least five  standard  solutions:  one should
      be near the detection limit; one should be near, but below, the expected
      concentrations of the analyte; one should be near, but  above, the expected
      concentrations  of  the  analyte.    The   range of  standard  solution
      concentrations used should  not  exceed the working range of the GC system.

            7.3.4 Prepare  a  calibration  curve  using  the  peak  areas of  the
      standards (retention time of acrylonitrile  under the  conditions of Sec.
      7.2 is approximately 2.3  minutes).    If the calibration curve deviates
      significantly from a straight line,  prepare a  new calibration curve with
      the existing standards, or,  prepare  new standards  and a new  calibration
      curve.   See  Method  8000,  Sec.   7.4.2,  for  additional   guidance  on
      calibration by the external  standard method.
                                   8031 - 4
                                Revision 0
                            September 1994

-------
      7.4   Sample Analysis

            7.4.1 Inject  3  ^L   of  the  sample  extract,   using   the  same
      chromatographic conditions  used  to  prepare the standard curve.  Calculate
      the concentration of acrylonitrile in the extract, using the area of the
      peak, against the calibration curve prepared in Sec. 7.3.4.


8.0   QUALITY CONTROL

      8.1   Refer to Chapter One and  Method  8000  for  specific quality control
procedures.

      8.2   Prior  to  preparation  of  stock  solutions,  methanol  and  methyl
tert-butyl ether reagents should  be analyzed gas chromatographically under the
conditions described in Sec. 7.2, to determine possible interferences with the
acrylonitrile peak.   If the solvent blanks show contamination,  a different batch
of solvents should be used.


9.0   METHOD PERFORMANCE

      9.1   Method 8031 was tested  in a single laboratory over  a period of days.
Duplicate samples and one spiked  sample were run for each calculation.  The GC
was calibrated daily.  Results are presented in Table 1.


10.0  REFERENCES

1.    K.L. Anderson,  "The  Determination  of Trace Amounts  of  Acrylonitrile in
      Water by Specific Nitrogen  Detector Gas Chromatograph",  American Cynamid
      Report No. WI-88-13, 1988.
                                   8031 - 5                         Revision 0
                                                                September 1994

-------
                       TABLE  1

         SINGLE LABORATORY METHOD PERFORMANCE
                  CONCENTRATION
  SAMPLE          SPIKE (ng/L)        % RECOVERY
      A                 60                100
      B                 60                105
      C                 40                 86
      D                 40                100
      E                 40                 88
      F                 60                 94

Average                                    96
                       8031 - 6                         Revision 0
                                                    September 1994

-------
            METHOD 8031
ACRYLONITRILE BY GAS CHROMATOGRAPHY
                 Start
7.1.1 Extract 40 mL
of sample with methyl
t-butyl ether in 90 mL
bottle.
>
1
7.2 Set
Chromatographic
conditions.
^
r
7.3.1 Flush GC
system with 30 uL
methyl t-butyl ether.
>
f
7.3.2 Analyze 3 uL
of sample blank.
^
r
7.3.3 - 7.3.4 Establish
calibration curve with
at least 5 stds.
^l
r
7.4 Sample Analysis
x
r
                   Stop
              8031 - 7
    Revision 0
September 1994

-------
00
o

-------
                                  METHOD  8032

                       ACRYLAMIDE BY GAS CHROMAT06RAPHY


1.0   SCOPE AND APPLICATION

      1.1   Method 8032 is  used to determine trace amounts of acrylamide monomer
in aqueous matrices.   This method may  be  applicable to other  matrices.   The
following compound can be determined by this method:
      Compound Name                             CAS No."


      Acrylamide                                79-06-01


      a  Chemical  Abstract  Services  Registry  Number.

      1.2   The method detection limit (MDL)  in clean water is 0.032

      1.3   This method is restricted to  use  by,  or  under the supervision of,
analysts  experienced  in  the  use  of gas  chromatographs  and  skilled  in  the
interpretation of gas  chromatograms.   Each  analyst must demonstrate the ability
to generate acceptable results with this method.


2.0   SUMMARY OF METHOD

      2.1   Method 8032 is based on bromination of the acrylamide double bond.
The reaction product  (2,3-dibromopropionamide)  is  extracted  from the reaction
mixture with ethyl acetate,  after salting  out with sodium  sulfate.  The extract
is cleaned up using a Florisil column, and analyzed by gas chromatography with
electron capture detection (GC/ECD).

      2.2   Compound  identification  should   be  supported  by   at  least  one
additional qualitative technique.  Analysis using a second gas chromatographic
column  or  gas  chromatography/mass  spectrometry  may  be used  for  compound
confirmation.


3.0   INTERFERENCES

      3.1   No interference is observed from sea water or in the presence of 8.0%
of ammonium  ions  derived  from  ammonium bromide.   Impurities from  potassium
bromide are removed by the Florisil  clean  up  procedure.
                                   8032 - 1                         Revision 0
                                                                September 1994

-------
4.0   APPARATUS AND MATERIALS

      4,1   Gas chromatographic System

            4.1.1 Gas chromatograph suitable for on-column injections with all
      required accessories,  including  detector,  analytical  columns,  recorder,
      gases, and syringes.  A data  system  for measuring peak heights and/or peak
      areas is recommended.

            4.1.2 Column:   2 m  x  3 mm glass column, 5%  FFAP  (free  fatty acid
      polyester) on 60-80 mesh acid washed Chromosorb W,  or equivalent.

            4.1.3 Detector:  electron capture detector.

      4.2   Kuderna-Danish  (K-D) apparatus.

            4.2.1 Concentrator tube -  10 mL graduated (Kontes K-570050-1025 or
      equivalent).  A  ground glass stopper is used  to prevent  evaporation of
      extracts.

            4.2.2 Evaporation  flask  -    500  ml  (Kontes    K-570001-500   or
      equivalent).   Attach to  concentrator  tube  with   springs,  clamps,  or
      equivalent.

            4.2.3 Snyder column  -   Three .ball macro  (Kontes  K-503000-0121 or
      equivalent).

            4.2.4 Snyder column  -   Two   ball  micro (Kontes  K-569001-0219 or
      equivalent).

            4.2.5 Springs -  1/2 inch (Kontes K-662750 or equivalent).

      4.3   Separatory funnel - 150 ml.

      4.4   Volumetric flask  (Class A)  -  100  ml,  with  ground  glass stopper;
25 ml, amber, with ground glass stopper.

      4.5   Syringe - 5 ml.

      4.6   Microsyringes - 5 ^i,  100 /^L.

      4.7   Pipets (Class A).

      4.8   Glass column (30 cm x 2 cm).

      4.9   Mechanical shaker.


5.0   REAGENTS

      5.1   Reagent grade chemicals  shall be used  in all tests.  Unless otherwise
indicated, it is intended that  all  reagents shall conform to the specifications
of the Committee on Analytical  Reagents of the  American Chemical Society, where


                                   8032 - 2                         Revision 0
                                                                September 1994

-------
such specifications  are  available.   Other grades may  be used, provided it  is
first ascertained that the reagent is of sufficiently high purity  to  permit  its
use without lessening the accuracy of the determination.

      5.2   Organic-free reagent water.   All  references to water in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.3   Solvents

            5.3.1 Ethyl acetate, C2H5C02C2H5.   Pesticide quality, or equivalent.

            5.3.2 Diethyl ether,  C2H5OC2H5.   Pesticide quality, or equivalent.
      Must  be  free  of  peroxides as  indicated by test  strips  (EM  Quant,   or
      equivalent).   Procedures  for  removal  of peroxides  are provided with  the
      test strips.  After cleanup, 20 ml of ethyl alcohol preservative must  be
      added to each liter of ether.

            5.3.3 Methanol, CH3OH.  Pesticide quality, or equivalent.

            5.3.4 Benzene, C6H6.   Pesticide  quality,  or equivalent.

            5.3.5 Acetone, CH3COCH3.   Pesticide quality,  or equivalent.

      5.4   Saturated bromine water.  Prepare by shaking organic-free reagent
water with bromine and allowing to stand for  1  hour,  in  the  dark, at  4°C.   Use
the aqueous phase.

      5.5   Sodium sulfate (anhydrous, granular), Na2S04.   Purify  by heating  at
400°C for 4 hours in a shallow tray, or by precleaning the sodium sulfate with
methylene chloride.   If the sodium sulfate is precleaned with methylene  chloride,
a method blank must be analyzed,  demonstrating  that there is no interference from
the sodium sulfate.

      5.6   Sodium thiosulfate, Na2S203,  1 M aqueous  solution.

      5.7   Potassium bromide, KBr, prepared  for infrared analysis.

      5.8   Concentrated hydrobromic acid, HBr, specific  gravity  1.48.

      5.9   Acrylamide  monomer,  H2C:CHCONH2,   electrophoresis  reagent  grade,
minimum 95% purity.

      5.10  Dimethyl phthalate, C6H4(COOCH3)2,  99.0%  purity.

      5.11  Florisil (60/100 mesh):   Prepare Florisil by activating at 130°C  for
at least 16 hours.  Alternatively, store Florisil in an  oven at 130°C.  Before
use, cool the Florisil in a desiccator.   Pack  5 g of  the  Florisil, suspended  in
benzene, in a glass column (Sec. 4.8).

      5.12  Stock standard solutions

            5.12.1      Prepare a stock  standard solution of  acrylamide monomer
      as specified in Sec. 5.12.1.1.   When  compound purity is assayed  to be  96%

                                   8032 - 3                         Revision 0
                                                                September 1994

-------
      or greater, the weight  can  be used without correction  to  calculate  the
      concentration  of the  stock standard.  Commercially prepared standards  can
      be used at  any concentration  if they are certified by the manufacturer or
      by an independent  source.

                  5.12.1.1     Dissolve  105.3  mg  of  acrylamide  monomer   in
            organic-free reagent water  in a 100 ml volumetric flask, and dilute
            to the mark  with organic-free reagent water. Dilute the solution of
            acrylamide monomer so  as  to obtain  standard  solutions containing
            0.1 - 10 mg/L of aery1 amide monomer.

      5.13  Calibration  standards

            5.13.1      Dilute the acrylamide stock solution with organic-free
      reagent water to  produce standard solutions  containing 0.1-5  mg/L of
      acrylamide.  Prior to injection the calibration standards are reacted  and
      extracted in the same manner as environmental  samples (Sec. 7).

      5.14  Internal standards

            5.14.1      The suggested internal  standard is dimethyl phthalate.
      Prepare  a  solution containing 100 mg/L  of dimethyl  phthalate  in  ethyl
      acetate.  The concentration  of dimethyl phthalate in the sample extracts
      and calibration standards should be 4 mg/L.


6.0   SAMPLE COLLECTION, PRESERVATION,  AND  HANDLING

      6.1   See the  introductory  material  to this chapter,  Organic  Analytes,
Sec. 4.1.
7.0   PROCEDURE

      7.1   Bromination

            7.1.1 Pipet 50 mL  of  sample into a 100 mL  glass  stoppered flask.
      Dissolve 7.5 g of potassium bromide into the sample,  with stirring.

            7.1.2 Adjust the pH of  the  solution  with  concentrated hydrobromic
      acid until  the pH is between 1 and 3.

            7.1.3 Wrap the flask with aluminum foil  in order to exclude light.
      Add 2.5 mL  of  saturated bromine water,  with  stirring.   Store the flask and
      contents in the dark, at 0°C,  for  at least  1 hour.

            7.1.4 After reacting the solution  for at  least an hour,  decompose
      the excess  of  bromine by adding 1 M sodium thiosulfate  solution, dropwise,
      until the color of the solution is discharged.

            7.1.5 Add 15 g of sodium sulfate, using a magnetic  stirrer to effect
      vigorous stirring.
                                   8032 - 4                         Revision 0
                                                                September 1994

-------
      7.2   Extraction

            7.2.1 Transfer the solution into  a  150 mL separatory funnel.  Rinse
      the reaction flask three times  with  1 ml  aliquots of organic-free reagent
      water.  Transfer the rinsings into the separatory funnel.

            7.2.2 Extract the aqueous solution  with  two 10 ml portions of ethyl
      acetate for 2 min each, using a mechanical shaker (240 strokes per min).
      Dry the organic phase with 1 g of sodium sulfate.

            7.2.3 Transfer  the  organic phase  into  a  25  ml  amber  volumetric
      flask.   Rinse  the  sodium  sulfate with  three  1.5  ml portions  of ethyl
      acetate and combine the rinsings with the organic phase.

            7.2.4 Add exactly 100 /ig  of dimethyl phthalate to the flask and make
      the solution  up to  the 25 ml  mark with  ethyl  acetate.   Inject 5  /LtL
      portions of this solution into the gas chromatograph.

      7.3   Florisil  cleanup:  Whenever interferences are observed, the samples
should be cleaned up as follows.

            7.3.1 Transfer the dried  extract  into a Kuderna-Danish  evaporator
      with  15  ml of  benzene.   Evaporate the  solvent  at  70°C under  reduced
      pressure, and concentrate the solution to about 3 ml.

            7.3.2 Add  50  ml of benzene and  subject the solution to  Florisil
      column chromatography at a flow rate of 3 mL/min.  Elute the column first
      with 50 ml of diethyl  ether/benzene  (1:4) at a flow rate of 5 mL/min,  and
      then with  25  ml of  acetone/benzene (2:1)  at a  flow rate of  2  mL/min.
      Discard all of the  first eluate and the initial  9 ml portion of the second
      eluate,  and use  the  remainder  for the determination,  using  dimethyl
      phthalate (4 mg/L) as  an internal standard.

            NOTE: Benzene  is  toxic,   and should  be  only be  used  under  a
                  ventilated laboratory hood.

      7.4   Gas chromatographic conditions:
      Nitrogen carrier gas flow rate:     40 mL/min
      Column temperature:                 165°C.
      Injector temperature:               180°C
      Detector temperature:               185°C.
      Injection volume:                   5 ^l

      7.5   Calibration:

            7.5.1 Inject  5  p.1 of  a  sample blank  (organic-free  reagent water
      carried through all sample storage, handling, bromination and extraction
      procedures).

            7.5.2 Prepare standard solutions of acrylamide  as described  in Sec.
      5.13.1.   Brominate and extract each  standard solution  as  described  in
      Sees. 7.1 and 7.2.


                                   8032  - 5                         Revision 0
                                                                September 1994

-------
            7.5.2.1     Inject 5 ^L of each of a minimum of five standard
      solutions: one should  be  near the detection  limit;  one should be
      near, but below,  the  expected concentrations of  the  analyte; one
      should  be near,   but  above,  the expected  concentrations  of the
      analyte.

            7.5.2.2     Prepare a calibration curve using the peak areas
      of the standards.  If the calibration curve deviates significantly
      from  a  straight  line,  prepare  a new  calibration curve  with the
      existing standards,  or,  prepare  new standards  and a new calibration
      curve.   See  Method  8000,  Sec.  7.4.3,  for  additional  guidance on
      calibration by the internal standard method.

            7.5.2.3     Calculate the  response factor  for  each standard
      according to Equation 1.

                  (Ps)  (MJ
            RF = 	              Equation  1
                  (Pis)  (MA)

            RF    =     Response factor
            Ps    =     Peak height of acrylamide
            Mis    =     Amount of internal  standard injected (ng)
            Pis    =     Peak height of internal standard
            MA    =     Amount of acrylamide injected (ng)

      7.5.3 Calculate the  mean response factor according to Equation 2.

                  n
                  ri


            RF = 	                        Equation 2
                   n

            RF    =     Mean response factor
            RF    =     Response    factors    from   standard   analyses
                        (calculated in Equation 1)
            n     =     Number of analyses

7.6   Gas chromatographic  analysis:

      7.6.1 Inject  5  ^l  portions  of each  sample  (containing  4  mg/L
internal  standard)   into  the  gas  chromatograph.   An  example  GC/ECD
chromatogram is shown in Figure 1.

      7.6.2 The concentration of  acrylamide monomer  in the sample is given
by Equation 3.
      [A] = - = -             Equation 3
             (Pis) (RF) (V,) (V.)
      [A]   =     Concentration of acrylamide monomer in sample (mg/L)

                             8032 - 6                         Revision 0
                                                          September 1994

-------
            PA    =     Peak height of acrylamide monomer
            Mjs    =     Amount of internal standard injected (ng)
            Vs    =     Total volume of sample (mL)
            P^    =     Peak height of internal standard
            RF    =     Mean response factor from Equation 2
            V|     =     Injection volume
8.0   QUALITY CONTROL

      8.1   Refer to Chapter One  and  Method  8000  for specific quality control
procedures.


9.0   METHOD PERFORMANCE

      9.1   The  following  performance  data  have  been  generated  under  the
conditions described in this method:

            9.1.1 The calibration curve  for Method 8032 is linear over the range
      0-5 M9/L of acrylamide monomer.

            9.1.2 The limit of  detection for  an aqueous solution is 0.032 M9/L.

            9.1.3 The yields of the  brominated compound are 85.2 +  3.3% and 83.3
      + 0.9%,  at  fortification concentrations of 1.0 and 5.0 M9/L. respectively.

      9.2   Table 1  provides the recoveries  of acrylamide  monomer  from river
water, sewage effluent, and sea water.

      9.3   The recovery of the bromination product as a  function of the amount
of  potassium  bromide  and  hydrobromic acid  added  to the  sample  is  shown  in
Figure 2.

      9.4   The effect of the reaction time on the recovery of the bromination
product is shown  in  Figure  3.  The yield was constant when the reaction time was
more than 1 hour.

      9.5   Figure 4 shows  the  recovery of the bromination product  as a function
of the initial  pH from 1 to 7.35.  The yield  was constant within this pH range.
The use of conventional buffer solutions, such as sodium acetate - acetic acid
solution or phosphate solution, caused a significant decrease in yield.


10.0  REFERENCES

1.    Hashimoto,   A.,   "Improved  Method  for  the  Determination  of  Acrylamide
      Monomer in Water by Means of Gas-Liquid Chromatography with  an Electron-
      capture Detector," Analyst, 101:932-938, 1976.
                                   8032 - 7                         Revision 0
                                                                September 1994

-------
                                                  TABLE  1

                                RECOVERY OF ACRYLAMIDE  FROM WATER SAMPLES AS
                                          2,3-DIBROMOPROPIONAMIDE
Sample
Matrix
Standard
River Water
Sewage
Effluent
Sea Water
Acryl amide
Monomer
Spiked//ng
0.05
0.20
0.25
0.20
0.20
0.20
Amount of 2
Calculated
0.162
0.649
0.812
0.649
0.649
0.649
,3-DBPAa/Mg
Found0
0.138
0.535
0.677
0.531
0.542
0.524
Overall
Bromination
Recovery
%b
85.2
82.4
83.3
81.8
83.5
80.7
Recovery of
Acryl amide
Monomer, %b
—
99.4
101.3
98.8
Coefficient
of
Variation
3.3
1.0
0.9
2.5
3.0
3.5
a  2,3-Dibromopropionamide

b  Mean of five replicate determinations
                                                  8032 - 8
    Revision 0
September 1994

-------
                                   Figure 1
                s
                JS
                                                         A


                                                         B
                                •   I    10   12   U

                                   Timt/min
Typical  gas  chromatograms of  the bromination  product obtained  from  aqueous
acrylamide monomer solution:

   A.    Untreated
   B.    With  Florisil  cleanup
   BL.   Chromatogram of blank, concentrated five-fold before gas chromatographic
        analysis.
Peaks:

   1.
   2.
   4-7.
2,3-Dibromopropionamide
Dimethyl phthalate
Impurities from potassium bromide
Sample size = 100 ml; acrylamide monomer = 0.1
                                   8032 - 9
                                                        Revision 0
                                                    September 1994

-------
                                  Figure 2
                    §
                    V
                    cc
                               5    10    IS    20    25
                             Amount of KBr/g ptr SO ml
                         i     . I      •     '     «     n
                         0     2    4     6     8    10

                            Amount of H8f/ml ptr 50 ml
Effect of  (A)  potassium  bromide  and (B)  hydrobromic  acid on  the  yield of
bromination.   Sample  size  « 50 ml; acrylamide monomer = 0.25 /*g
                                  8032  -  10
    Revision 0
September 1994

-------
                                   Figure 3
                                                     24
Effect of reaction time on the bromination.   Reaction conditions:

   50 ml of sample;
   0.25 ng of acrylamide monomer;
   7.5 g of potassium bromide;
   2.5 ml of saturated bromine water

Extraction conditions:

   15 g of sodium sulfate;
   extraction at pH  2;
   solvent = 10 ml of ethyl  acetate (X2)
                                  8032  -  11
    Revision 0
September 1994

-------
                                   Figure 4
                   100
                    50
                     01    2    349671

                                       PM
Effect of initial pH on the  bromination.  Reaction and extraction conditions as
in Figure 3.   The pH was adjusted to below 3 with concentrated hydrobromic acid,
and to  4-5 with  dilute hydrobromic acid.   Reaction  at pH 6 was  in distilled
water,  pH  7.35 was  achieved  by careful addition  of dilute sodium hydroxide
solution. The broken line shows the result obtained by the  use of sodium acetate
-  acetic acid buffer solution.
                                  8032  -  12
    Revision 0
September 1994

-------
                           METHOD 8032
             ACRYLAMIOE  BY  GAS  CHROMATOGRAPHY
Start
T
7.1 Bromination
1
7.1 .1 Dissolve 7.5 g KBr into
50 ml sample in flask.
i
7. 1.2 Adjust soln. pHwith
concentrated HBr to between
1 and 3.
1
7. 1.3 Wrap soln. flask with
aluminum. Add 2.5 ml satd.
bromine water, stir, store at
0 C for 1 hr.
1
7. 1.4 Add 1 M sodium
thiosulfate dropwise to flask to
decompose excess bromine.
I
7. 1. 5 Add 15 g sodium
sulfate, and stir.



1
7.2 Extraction
i
7.2.1 Transfer flask soln. to
sap. funnel along with rinses.
i
7.2.2 Extract soln. twice w/ethyl
acetate. Dry organic phase
using sodium sulfate.
1
7.2.3 Transfer organic phase
and rinses into amber
glass flask.
1
7.2.4 Add 100 ug dimethyl
phthalate to flask, dilute to
mark Inject 5 uL into GC.
1
7.3 Florisil Cleanup
1
7.3.1 Transfer dried extract to
K-D assembly w/benzene
Concentrate to 3 ml at 70 C
under reduced pressure.
                                                      7.3.2 Add 50 ml benzene to
                                                      solution. Pass soln. through
                                                      Florisil column.  Elute with
                                                      diethyl ether/benzene, then
                                                      acetone/benzene.  Collect
                                                      the second elution train (less
                                                      initial 9 ml) for analysis.
                             8032  -  13
     Revision  0
September  1994

-------
          METHOD  8032
            continued
      7.4 GC Conditions
       7.5 Calibration
 7.5.1 Inject 5 uL sample blank.
7.5.2 Brominate and extract std.
solns. similar to the samples.
. 1 Inject 5 uL of each of the
  minimum 5 stds.
.2 Plot peak are vs. [).
.3 Calculate response factor
  (RF) for each [ ].
 7.5.3 Calculate mean RF from
 eqn. 2.
            I
7.6 GC Analysis
1

7.6.1 Inject 5 uL sample containing
internal std. into GC.
\

7.6.2 Calculate acrylamide monomer
concentration in sample using
eqn. 3.
            8032  -  14
     Revision  0
September  1994

-------
00
o

-------
                                 METHOD 8040A

                         PHENOLS  BY GAS  CHROMATOGRAPHY
1.0   SCOPE AND APPLICATION

      1.1   Method  8040  is  used  to  determine the  concentration of  various
phenolic compounds.  The following compounds can be determined by this method:
Compound Name
               Appropriate Technique

  CAS No.a   3510   3520    3540  3550  3580
2-sec-Butyl-4,6-dinitrophenol
  (DNBP, Dinoseb)
4-Chloro-3-methylphenol
2-Chlorophenol
Cresols (methyl phenols)
2-Cyclohexyl-4,6-dinitrophenol
2,4-Dichlorophenol
2,6-Dichlorophenol
2,4-Dimethylphenol
2,4-Dinitrophenol
2-Methyl-4,6-dinitrophenol
2-Nitropheno]
4-Nitrophenol
Pentachlorophenol
Phenol
Tetrachlorophenols
Trichlorophenols
2,4,6-Trichlorophenol
   88-85-7   X      NO     NO    NO    X

   -59-50-7   X      X      X     X     X
   95-57-8   X      X      X     X     X
 1319-77-3   X      ND     NO    ND    X
  131-89-5   X      ND     ND    ND    LR
  120-83-2   X      X      X     X     X
   87-65-0   X      ND     ND    ND    X
  105-67-9   X      X      X     X     X
   51-28-5   X      X      X     X     X
  534-52-1   X      X      X     X     X
   88-75-5   X      X      X     X     X
  100-02-7   X      X      X     X     X
   87-86-5   X      X      X     X     X
  108-95-2 DC(28)   X      X     X     X
25167-83-3   X      ND     ND    ND    X
25167-82-2   X      X      X     X     X
   88-06-2   X      X      X     X     X
a      Chemical Abstract Services Registry Number.
DC =   Unfavorable distribution coefficient  (number  in  parenthesis is percent
       recovery).
LR =   Low response.
ND =   Not determined.
X  =   Greater than 70 percent recovery by this technique.

       1.2  Table 1 lists the method detection limit  for the target analytes in
water.  Table 2 lists the estimated quantitation limit (EQL) for all  matrices.
2.0    SUMMARY OF METHOD

       2.1  Method 8040 provides gas chromatographic conditions for the detection
of  phenolic  compounds.   Prior  to  analysis,  samples  must be  extracted using
appropriate techniques (see Chapter Two  for guidance).   Both neat and diluted
organic  liquids  (Method  3580,  Waste  Dilution)  may  be  analyzed by  direct
                                   8040A - 1
                                 Revision 1
                                  July 1992

-------
injection.  A 2 to 5 p.1  sample  is  injected  into a gas chromatograph using the
solvent flush technique,  and compounds in the GC effluent are detected fay a flame
ionization detector (FID).

       2.2  Method 8040 also  provides  for the preparation of pentafluorobenzyl-
bromide  (PFB)  derivatives,  with  additional cleanup  procedures  for electron
capture gas  chromatography.   This  is to lower  the  detection  limits  of some
phenols and to aid the analyst in the elimination of interferences.


3.0    INTERFERENCES

       3.1  Refer to Methods  3500,  3600, and 8000.

       3.2  Solvents, reagents,  glassware, and other sample processing hardware
may yield discrete artifacts and/or elevated  baselines causing misinterpretation
of gas chromatograms.  All of these materials must be demonstrated to be free
from interferences, under the conditions of the analysis, by analyzing reagent
blanks.    Specific  selection  of  reagents  and  purification  of  solvents  by
distillation in all-glass systems may be required.

       3.3  Interferences coextracted from samples will vary considerably from
source to  source, depending  upon  the waste being sampled.   Although general
cleanup techniques are recommended as part  of this method,  unique samples may
require additional cleanup.

       3.4  The decomposition of some  analytes under basic extraction conditions
has been demonstrated.  Specifically,  phenols may react to form tannates.  These
reactions increase with increasing  pH, and are decreased by the  shorter reaction
times available in Method 3510.

       3.5  The flame  ionization  detector (FID)  is very susceptible to false
positives caused by the presence of hydrocarbons  commonly found  in  samples from
waste sites.   The  problem may be minimized by applying  acid-base  cleanup (Method
3650) and/or  alumina column chromatography (Method 3611) prior to GC/FID analysis
or  using  the derivatization  technique  and  analyzing by  GC/electron capture
detector.  Initial site investigation  should always be  performed utilizing GC/MS
analysis to  characterize the site and  determine  the  feasibility of utilizing
Method 8040 with  a GC/FID.


4.0    APPARATUS AND MATERIALS

       4.1  Gas chromatograph

            4.1.1 Gas  Chromatograph   -   Analytical  system   complete  with  gas
       chromatograph  suitable  for   on-column   injections   and  all  required
       accessories, including detectors, column supplies, recorder, gases, and
       syringes.  A data system  for measuring peak areas  and/or  peak  heights is
       recommended.

            4.1.2 Columns

                  4.1.2.1     Column  for underivatized phenols - 1.8 m x  2.0 mm

                                   8040A - 2                         Revision  1
                                                                     July 1992

-------
            ID glass column packed with 1% SP-1240DA on Supelcoport 80/100 mesh,
            or equivalent.

                  4.1.2.2     Column for derivatized phenols - 1.8 m x 2 mm ID
            glass column  packed  with  5% OV-17 on  Chromosorb  W-AW-DMCS 80/100
            mesh, or equivalent.

            4.1.3 Detectors  -  Flame ionization (FID) and electron capture (ECD).

       4.2  Reaction vial -  20 ml,  with Teflon lined screw-cap or crimp top.

       4.3  Volumetric flask,  Class  A  -  Appropriate sizes with  ground-glass
stoppers.

       4.4  Kuderna-Danish (K-D)  apparatus

            4.4.1 Concentrator tube -  10 ml, graduated (Kontes K-570050-1025 or
       equivalent).    Ground-glass  stopper is  used to prevent  evaporation of
       extracts.

            4.4.2 Evaporation  flask   -  500  ml   (Kontes   K-570001-500   or
       equivalent).    Attach  to concentrator tube  with  springs,  clamps  or
       equivalent.

            4.4.3 Snyder  column  -  Three  ball  macro (Kontes  K-503000-0121 or
       equivalent).

            4.4.4 Snyder  column  -  Two ball  micro  (Kontes  K-569001-0219  or
       equivalent).

            4.4.5 Springs -  1/2 inch (Kontes K-662750 or equivalent).

       4.5  Boiling chips - Solvent extracted,  approximately 10/40 mesh  (silicon
carbide or equivalent).

       4.6  Water  bath  - Heated,  with  concentric  ring  cover,   capable  of
temperature control   (± 5°C).   The bath should  be  used in  a hood.

       4.7  Microsyringe  - 10 u.L.

       4.8  Syringe  - 5 ml.

       4.9  Balance  - analytical, 0.0001 g..


5.0    REAGENTS

       5.1  Reagent grade chemicals shall  be used in all tests. Unless otherwise
indicated, it is intended that all  reagents  shall conform to the specifications
of the Committee on  Analytical Reagents of the American Chemical Society, where
such specifications are available. Other grades may  be used, provided it is first
ascertained that  the reagent  is  of sufficiently  high purity to permit its use
without  lessening the accuracy of the  determination.


                                   8040A - 3                         Revision 1
                                                                     July  1992

-------
       5.2  Organic-free reagent water -  All references to water in this method
refer to organic-free reagent water, as defined in Chapter One.

       5.3  Hexane, CH3(CH2)4CH3 - Pesticide quality or equivalent.

       5.4  2-Propanol, (CH3)2CHOH  - Pesticide  quality or equivalent.

       5.5  Toluene, C6H5CH3 - Pesticide quality or equivalent.

       5.6  Derivatization reagent - Add  1  ml pentafluorobenzyl bromide and 1 g
18-crown-6-ether  to  a 50  mi  volumetric  flask  and  dilute to  volume  with
2-propanol.  Prepare  fresh weekly.  This operation  should  be carried out in a
hood.  Store at 4°C and protect from light.

            5.6.1 Pentafluorobenzyl  bromide   (alpha-Bromopentafluorotoluene),
       C6F5CH2Br.  97% minimum purity.

            NOTE:  This chemical is a lachrymator.

            5.6.2 18-crown-6-ether  (1,4,7,10,13,16-Hexaoxacyclooctadecane)  -
       98% minimum purity.

            NOTE;  This chemical is highly toxic.

       5.7  Potassium carbonate  (Powdered)., K2C03.

       5.8  Stock standard solutions

            5.8.1 Prepare  stock  standard  solution  at  a  concentration  of
       1000 mg/L  by  dissolving  0.0100  g  of  assayed  reference material  in
       2-propanol and diluting to  volume in a  10 ml volumetric  flask.  Larger
       volumes can  be used  at the convenience of  the  analyst.   When compound
       purity is assayed  to be  96% or greater, the weight can be used without
       correction  to  calculate  the   concentration  of  the  stock  standard.
       Commercially prepared stock standards can be used at any concentration if
       they are certified by the manufacturer or by an independent source.

            5.8.2 Transfer the stock standard solutions into bottles with Teflon
       lined screw-caps or  crimp tops.  Store at  4°C  and protect from light.
       Stock standards should be checked  frequently for signs of degradation or
       evaporation,  especially  just prior to  preparing  calibration standards
       from them.

            5.8.3 Stock standard solutions must be  replaced after one year, or
       sooner if comparison with check standards indicates a problem.

       5.9  Calibration standards -  Prepare calibration  standards at a minimum
of five concentrations through dilution of  the  stock standards with 2-propanol.
One  of the concentrations should  be  at  a concentration  near,  but  above, the
method detection limit.   The  remaining concentrations should correspond to the
expected  range  of  concentrations  found  in real  samples  or should  define the
working  range  of the GC.  Calibration  solutions  must  be  replaced  after six
months, or sooner,  if comparison with check standards indicates a problem.


                                   8040A -  4                         Revision  1
                                                                     July  1992

-------
       5.10 Internal standards (if internal standard calibration is used)  - To
use this approach, the analyst must select one or more internal standards  that
are similar in analytical  behavior to  the  compounds of interest.   The analyst
must further demonstrate that the measurement  of the internal  standard is not
affected by method or matrix  interferences.   Because of these limitations, no
internal standard can be suggested that is applicable to all samples.

            5.10.1      Prepare  calibration  standards  at  a  minimum  of  five
       concentrations for each analyte as described in Section 5.9.

            5.10.2      To  each  calibration  standard,  add a  known  constant
       amount of one or more  internal  standards,  and dilute to volume with 2-
       propanol.

            5.10.3      Analyze each calibration standard according to Section
       7.0.

       5.11 Surrogate standards - The  analyst  should monitor the performance of
the  extraction,   cleanup  (if  necessary),  and  analytical  system  and   the
effectiveness of the method in dealing with each sample matrix by spiking  each
sample, standard, and organic-free reagent  water  blank with phenolic surrogates
(e.g.  2-fluorophenol  and 2,4,6-tribromophenol)  recommended to encompass the range
of the temperature program used in this method.   Method 3500 details instructions
on the preparation of acid surrogates.  Deuterated analogs of analytes should not
be used as  surrogates for gas chromatographic analysis due to  coelution problems.


6.0    SAMPLE COLLECTION, PRESERVATION, AND HANDLING

       6.1   See  the  introductory material  to  this  chapter,  Organic Analytes,
Section 4.1.  Extracts must be stored under refrigeration and analyzed within 40
days of extraction.


7.0    PROCEDURE

       7.1   Extraction

            7.1.1 Refer to Chapter Two for  guidance  on choosing the appropriate
       extraction procedure.   In  general, water samples are extracted at a pH of
       less than  or  equal to 2 with methylene chloride,  using either Method 3510
       or 3520.   Solid samples are extracted using either Method 3540 or 3550,
       and  non-aqueous  samples  using Method  3580.   Extracts obtained  from
       application  of  either  Method  3540  or 3550  should undergo  Acid-Base
       Partition  Cleanup, using Method 3650.

            7.1.2 Prior to gas chromatographic analysis, the extraction solvent
       must be exchanged to 2-propanol.  The exchange is performed as follows:

                  7.1.2.1     Following concentration  of  the extract  to  1 mL
            using the macro-Snyder column,  allow the apparatus to cool and drain
            for  at least 10 minutes.
                                   8040A -  5                        Revision 1
                                                                     July 1992

-------
                  7.1.2.2     Remove the micro-Snyder column and rinse its lower
            joint  into  the  concentrator tube  with a  minimum  amount  of  2-
            propanol.   Adjust  the  extract  volume  to  1.0 ml.   Stopper  the
            concentrator  tube  and  store  refrigerated  at  4°C  if  further
            processing will  not be performed immediately.   If the extract will
            be stored longer than two days,  it should be transferred to a vial
            with a Teflon  lined  screw-cap or crimp top.   If the extract requires
            no   further  derivatization   or   cleanup,   proceed   with   gas
            chromatographic analysis.

       7.2  Gas chromatographic conditions (Recommended)

            7.2.1 Column for underivatized phenols -

            Carrier gas (N2)  flow rate:    30 ml/min
            Initial temperature:          80°C
            Temperature program:          80°C to 150°C  at  8°C/min
            Final Temperature:            150°C,  hold until all compounds have
                                          eluted.

            7.2.2 Column for derivatized phenols -

            Carrier gas (5% methane/95% argon)
            flow rate:                          30 mL/min
            Initial temperature:                200°C
            Temperature program:                isothermal,  hold   until   all
                                                compounds have eluted.

       7.3  Calibration -  Refer to Method 8000 for proper  calibration techniques.
Use Table 1 and especially Table 2 for guidance on selecting the lowest point on
the calibration curve.

            7.3.1 The procedure for  internal or external  calibration  may be used
       for the underivatized  phenols.  Refer to Method 8000  for  a description of
       each of these procedures.  If derivatization of the phenols is required,
       the method of  external calibration should  be  used  by injecting  five or
       more concentrations of calibration standards that  have also undergone
       derivatization and cleanup prior to instrument calibration.

       7.4  Gas chromatographic analysis

            7.4.1 Refer to Method 8000.   If the internal  standard calibration
       technique is used,  add 10 jiL of internal  standard to the sample prior to
       injection.

            7.4.2 Phenols are to be determined on a gas chromatograph equipped
       with a flame ionization detector according to the conditions listed for
       the 1% SP-1240DA column  (Section  7.2.1).   Table  1  summarizes estimated
       retention times and sensitivities that should  be  achieved  by  this method
       for  clean  water   samples.    Estimated  quantitation  limits   for  other
       matrices are list  in Table 2.

            7.4.3 Method  8000  provides  instructions  on  the analysis sequence,
       appropriate  dilutions,  establishing  daily retention  time windows, and

                                   8040A - 6                        Revision  1
                                                                     July  1992

-------
Identification criteria.  Include a mid-concentration standard after each
group of 10 samples in the analysis sequence.

     7.4.4 An  example  of a GC/FID chromatogram  for certain  phenols is
shown in  Figure  1.   Other packed  or  capillary (open-tubular)  columns,
chromatographic conditions, or detectors may  be used if the requirements
of Section 8.2 are met.

     7.4.5 Record the sample volume injected and the resulting peak sizes
(in area units or peak heights).

     7.4.6 Using either  the internal  or external calibration procedure
(Method 8000),  determine  the identity and quantity of each component peak
in the sample  chromatogram which  corresponds to  the compounds  used for
calibration purposes.  See Method 8000 for calculation equations.

     7.4.7 If  peak detection  using the  SP-1240DA column  with the flame
ionization detector is  prevented by interferences, PFB derivatives of the
phenols  should be  analyzed  on  a  gas  chromatograph equipped with  an
electron capture detector according to the conditions listed for the 5%
OV-17 column (Section 7.2.2).   The derivatization and cleanup procedure
is outlined in Sections  7.5 through 7.6.   Table  3 summarizes estimated
retention times for derivatives of some phenols using the conditions of
this method.

     7.4.8 Figure 2  shows  a  GC/ECD chromatogram  of PFB  derivatives of
certain phenols.

     7.4.9 Record the sample volume injected and the resulting peak sizes
(in area units or peak heights).

     7.4.10      Determine the  identity  and  quantity of  each component
peak in the sample chromatogram which corresponds to the compounds used
for calibration purposes.  The method of external calibration should be
used (see Method  8000 for guidance).  The concentration of the individual
compounds in the sample  is calculated as follows:


     Concentration (/ig/L)  =
where:
     A     »     Mass of underivatized phenol represented by area of peak
                 in  sample chromatogram,  determined  from calibration
                 curve (see Method 8000), ng.

     Vt    -     Total amount of column eluate or combined fractions from
                 which V,. was taken,  ML.

     B     =     Total volume of hexane added in Section 7.5.5, ml.

     D     -     Total   volume   of   2-propanol   extract  prior   to
                 derivatization, ml.

                            8040A -  7                        Revision 1
                                                              July 1992

-------
            V,    *     Volume injected, /nL.

            X     *     Volume of water extracted, ml, or weight of nonaqueous
                        sample extracted, g, from Section 7.1.  Either the dry
                        or wet  weight  of the  nonaqueous sample may  be  used,
                        depending upon the specific application of the data.

            C     »     Volume of hexane sample solution added to cleanup column
                        (Method 3630), ml.

            E     =     Volume   of   2-propanol   extract   carried   through
                        derivatization in Section 7.5.1, ml.

       7.5  Derivatization - If interferences prevent measurement of peak area
during  analysis  of the  extract by  flame  ionization gas  chromatography,  the
phenols must be derivatized and analyzed by electron capture gas chromatography.

            7.5.1 Pipet  a 1.0  ml aliquot  of the  2-propanol  stock  standard
       solution or of  the  sample extract into a glass  reaction vial.  Add 1.0 ml
       derivatization   reagent   (Section  5.3).    This  amount  of  reagent  is
       sufficient to derivatize  a solution whose total  phenolic content does not
       exceed 300 mg/L.

            7.5.2 Add  approximately  0.003  g  of  potassium  carbonate to  the
       solution and shake gently.

            7.5.3 Cap the mixture and heat it for 4 hours at 80°C in a hot water
       bath.

            7.5.4 Remove the solution from  the hot water bath and  allow it to
       cool.

            7.5.5 Add 10 mi hexane to the reaction vial and shake vigorously for
       1 minute.   Add  3.0 ml organic-free reagent water to the reaction vial and
       shake for 2 minutes.

            7.5.6 Decant the  organic  layer into a concentrator tube and cap with
       a glass stopper.  Proceed with cleanup procedure.

       7.6  Cleanup

            7.6.1 Cleanup of the derivatized extracts takes place using Method
       3630 (Silica Gel Cleanup),  in  which specific instructions for cleanup of
       the derivatized phenols  appear.

            7.6.2 Following column cleanup,  analyze the samples using GC/ECD, as
       described starting in Section  7.4.7.


8.0    QUALITY CONTROL

       8.1  Refer  to   Chapter  One for  specific  quality control  procedures.
Quality control  to validate  sample extraction is covered in Method 3500 and in


                                  8040A - 8                        Revision 1
                                                                     July 1992

-------
the extraction method used.   If extract  cleanup was performed, follow the QC in
Method 3600 and in the specific cleanup method.

       8.2  Procedures to check the GC system operation are found  in Method 8000,
Section 8.6.

            8.2.1 The quality  control  check sample concentrate  (Method  8000,
       Section 8.6)  should contain each analyte  of interest at a concentration
       of 100 mg/L in 2-propanol.

            8.2.2 Table 4 indicates the calibration and QC acceptance criteria
       for  this  method.    Table  5  gives  method  accuracy  and precision  as
       functions  of  concentration  for the analytes.  The contents of both tables
       should be  used  to evaluate a laboratory's  ability to perform and generate
       acceptable data by this method.

       8.3  Calculate surrogate standard  recovery  on  all  samples,  blanks,  and
spikes.   Determine  if the  recovery is within  limits  (limits  established  by
performing QC procedures outlined in Method 8000, Section 8.10).

            8.3.1 If recovery  is not within limits, the following is required.

            •     Check to  be  sure  that there are no  errors  in calculations,
                  surrogate  solutions  and  internal  standards.   Also,  check
                  instrument performance.

            •     Recalculate  the data  and/or reanalyze the  extract  if any of
                  the above checks reveal  a problem.

            •     Reextract and reanalyze  the  sample if none of the above are a
                  problem or flag the data as "estimated concentration."


9.0    METHOD PERFORMANCE

       9.1  The method was tested  by 20  laboratories using organic-free reagent
water, drinking water, surface  water, and three industrial wastewaters spiked at
six concentrations over the  range 12 to 450 M9/L-   Single operator precision,
overall precision, and method accuracy  were found to be  directly related to the
concentration of the analyte and essentially independent of the sample matrix.
Linear equations  to  describe these relationships  for a  flame ionization detector
are presented in Table 5.

       9.2  The accuracy and precision obtained  will be affected by the sample
matrix, sample-preparation technique, and calibration  procedures used.


10.0   REFERENCES

1.     Development and Application of Test Procedures for Specific Organic Toxic
       Substances in  Wastewaters.   Category 3 -  Chlorinated  Hydrocarbons  and
       Category  8  -  Phenols.    Report  for  EPA  Contract  68-03-2625  (in
       preparation).


                                  804-OA - 9                        Revision 1
                                                                     July 1992

-------
U.S. EPA 40  CFR  Part  136,  "Guidelines Establishing Test Procedures for
the Analysis  of  Pollutants Under  the Clean Water Act;  Final  Rule and
Interim Final Rule and Proposed Rule," October 26, 1984.

"Determination of Phenols  in  Industrial  and  Municipal  Wastewaters,"
Report for EPA Contract 68-03-2625 (in preparation).

"EPA Method Validation Study Test Method 604 (Phenols)," Report for EPA
Contract 68-03-2625 (in preparation).

Kawahara,  F.K.  "Microdetermination  of  Derivatives  of  Phenols  and
Mercaptans by Means of Electron Capture Gas Chromatography," Analytical
Chemistry, 40, 1009, 1968.

Burke, J.A.  "Gas Chromatography  for Pesticide  Residue  Analysis; Some
Practical Aspects," Journal  of the  Association  of Official  Analytical
Chemists, 48, 1037, 1965.
                           8040A  -  10                         Revision  1
                                                               July 1992

-------
                                   TABLE 1.
                FLAME IONIZATION GAS  CHROMATOGRAPHY OF PHENOLS*
                                                                 Method
                                             Retention  time       Detection
Analyte                                      (minutes)            limit
2-sec-Butyl -4,6-dinitrophenol (DNBP)
4-Chloro-3-methyl phenol                       7.50                   0.36
2-Chlorophenol                                1.70                   0.31
Cresols (methyl phenols)
2-Cyclohexyl -4,6-dinitrophenol
2,4-Dichlorophenol                            4.30                   0.39
2,6-Dichlorophenol
2,4-Dimethylphenol                            4.03                   0.32
2,4-Dinitrophenol                            10.00                  13.0
2-Methyl -4,6-dinitrophenol                   10.24                  16.0
2-Nitrophenol                                 2.00                   0.45
4-Nitrophenol                                24.25                   2.8
Pentachlorophenol                            12.42                   7.4
Phenol                                        3.01                   0.14
Tetrachl orophenol s
Trichlorophenols
2,4,6-Trichlorophenol                         6.05                   0.64


a -  1% SP-1240DA on Supelcoport 80/100 mesh column.
                                   TABLE 2.
                    DETERMINATION  OF  ESTIMATED QUANTITATION
                      LIMITS (EQLs)  FOR VARIOUS MATRICES"
   Matrix                                                   Factor6


Ground water                                                     10
Low-concentration soil by sonication with GPC cleanup           670
High-concentration soil and sludges by sonication            10,000
Non-water miscible waste                                    100,000


a  Sample EQLs are highly matrix-dependent.  The EQLs listed herein  are provided
   for guidance and may not always be achievable.

b  EQL =  [Method  detection limit (Table  1)]  X [Factor (Table  2)].   For non-
   aqueous samples,  the factor is on a wet-weight basis.
                                  8040A - 11                        Revision 1
                                                                     July 1992

-------
                                 TABLE 3.
         ELECTRON CAPTURE GAS CHROMATOGRAPHY OF PFB DERIVATIVES8


Parent compound
4-Chl oro-2-methyl phenol
2-Chlorophenol
2,4-Dichlorophenol
2, 4-Dimethyl phenol
2,4-Dinitrophenol
2-Methyl-4,6-dinitrophenol
2-Nitrophenol
4-Nitrophenol
Pentachlorophenol
Phenol
2,4,6-Trichlorophenol
Retention
time
(min)
4.8
3.3
5.8
2.9
46.9
36.6
9.1
14.0
28.8
1.8
7.0
Method
detection
limit (jug/L)
1.8
0.58
0.68
0.63


0.77
0.70
0.59
2.2
0.58
- 5% OV-17 on Chromosorb W-AW-OMCS 80/100 mesh column,
                                 8040A -  12                         Revision  1
                                                                    July 1992

-------
                                   TABLE 4.
                           QC ACCEPTANCE CRITERIA'


Analyte
4-Chl oro-3-methyl phenol
2-Chlorophenol
2,4-Dichlorophenol
2,4-Dimethylphenol
4, 6-Dinitro-2-methyl phenol
2,4-Dinitrophenol
2-Nitrophenol
4-Nitrophenol
Pentachlorophenol
Phenol
2,4,6-Trichlorophenol
s - Standard deviation of
x =• Average recovery for
Test
cone.
(M9/L)
100
100
100
100
100
100
100
100
100
100
100
four recovery
four recovery
Limit
for s
(M9/L)
16.6
27.0
25.1
33.3
25.0
36.0
22.5
19.0
32.4
14.1
16.6
measurements
measurements,
Range
for x
(M9/L)
56.7-113.4
54.1-110.2
59.7-103.3
50.4-100.0
42.4-123.6
31.7-125.1
56.6-103.8
22.7-100.0
56.7-113.5
32.4-100.0
60.8-110.4
, in M9/L.
in M9/L.
Recovery
Range
(%)
99-122
38-126
44-119
24-118
30-136
12-145
43-117
13-110
36-134
23-108
53-119


a   Criteria from 40  CFR Part 136 for Method  604.   These criteria  are  based
    directly upon the  method  performance data in Table 5.  Where necessary,  the
    limits for  recovery have  been  broadened to  assure applicability of  the
    limits to concentrations  below those  used to develop Table 5.
                                  8040A - 13                        Revision 1
                                                                     July 1992

-------
                                   TABLE 5.
         METHOD ACCURACY AND PRECISION AS FUNCTIONS OF CONCENTRATION*
Analyte
4-Chl oro-3-methyl phenol
2-Chlorophenol
2,4-Dichlorophenol
2,4-Dimethylphenol
4, 6-Dinitro-2-methyl phenol
2,4-Dinitrophenol
2-Nitrophenol
4-Nitrophenol
Pentachlorophenol
Phenol
2,4,6-Trichlorophenol
Accuracy, as
recovery, x'
(M9/U
0.87C-1.97
0.83C-0.84
0.81C+0.48
0.62C-1.64
0.84C-1.01
0.80C-1.58
0.81C-0.76
0.46C+0.18
0.83C+2.07
0.43C+0.11
0.86C-0.40
Single analyst
precision, s '
(M9/L)
O.llx-0.21
O.lSx+0.20
0.17X-0.02
0.30X-0.89
O.lBx+1.25
0.27X-1.15
0.15X+0.44
0.17X+2.43
0.22X-0.58
0.20X-0.88
O.lOx+0.53
Overall
precision,
S' (M9/L)
0.16X+1.41
0.21X+0.75
0.18X+0.62
0.25X+0.48
0.19X+5.85
0.29X+4.51
0.14x+3.84
0.19X+4.79
0.23X+0.57
0.17X+0.77
0.13X+2.40
X'
C

X
Expected  recovery  for  one  or  more  measurements  of  a  sample
containing a concentration of C,  in /ug/L.

Expected single  analyst  standard deviation of measurements  at an
average concentration of x, in /ig/L.

Expected interlaboratory standard  deviation  of  measurements  at an
average concentration found of x, in /ig/L.

True value for the concentration, in jug/l.

Average recovery found for  measurements  of samples  containing a
concentration of C, in /jg/L.
"From 40 CFR Part 136 for Method 604.
                                  8040A - 14
                                                        Revision  1
                                                         July  1992

-------
           Figure 1
 Gas  Chromatogram  of Phenols
   Column: 1% Sf-12*00A on Supdcooort
   Program: 80°C 0 MinutM 8°/Minute to 150°C
   0«ttcxer: Fl«m« formation
                                1
                                i
I      12      18      20      24
 RCTINT1ON TIME (MINUTES)
21
          8040A  -  15
        Revision 1
         July 1992

-------
                 Figure 2
Gas  Chromatogram of PFB Derivatives  of Phenols
              Column: 5% OV-17 on Giromaiorb W-AW
              T«mpof«unj: 200°C
              Octoetor: Iloetroft Cwtvrc
                                         A_
              «     «    »     24

           ftlTfNTtON TIMI (MINUTQ)
2t
31
                  8040A - 16
  Revision 1
   July 1992

-------
                       METHOD 8040A
             PHENOLS  BY  GAS  CHROMATOGRAPHY
 7  1 1 Cho
 appr opr i
 ex t r ac t1
 me thod {r
 to  Chapte
se
I*
n
f.r
 21
     7 1 2
   Enchange
  ex traction
  jo 1ven t Lo
  2-propjnol
  ^  2 Set gaj
chroma tography
  condition*
 7  3  Refar to
  Method 8000
  for propar
  calibration
  technique!
7  3  1  Inject at
    lea.t S
concen t ra11oni
of ca11bra 11on
   >tandard>
                 7 4 Perform
                 CC analy»ia
                 (see He thod
                    8000)
                  7 4 analyze
                using CC/FIO
                        8040A  -  17
                                                         Revision  1
                                                          July  1992

-------
          METHOD  8040A
           (Continued)
                           7 S Pr«p»r«
                           d«rivativ«t

No
749 Record
J*mpl* volum*
in j«ct*d and

                           1 6 Cleanup
                          using Method
                             3630
    7 1 10
IdvnIitify and
quantital« each
componvnl pvalc
7  4  7 Analyi.
    PFB
 d«riva tiv«a
ujing CC/ECD
    7 4 10
   Calculate
 conc«nt ration
    Stop
            8040A -  18
                            Revision 1
                             July  1992

-------
00
o

-------
                                  METHOD  8061

               PHTHALATE ESTERS BY CAPILLARY GAS CHROMATOGRAPHY
                   WITH ELECTRON CAPTURE DETECTION (GC/ECD)
1.0   SCOPE AND APPLICATION

      1.1   Method 8061 is used to determine the identities and concentrations
of various phthalate esters  in liquid, solid and sludge matrices.  The following
compounds can be determined by this method:
      Compound Name                                           CAS No.8


      Benzyl benzoate  (I.S.)                                 120-51-4
      Bis(2-ethylhexyl) phthalate                             117-81-7
      Butyl benzyl phthalate                                   85-68-7
      Di-n-butyl phthalate                                     84-74-2
      Diethyl phthalate                                        84-66-2
      Dimethyl phthalate                                      131-11-3
      Di-n-octyl phthalate                                    117-84-0


      8  Chemical  Abstract Services  Registry Number.

      1.2   Table  1  lists the  method  detection  limits  (MDL)  for  the target
analytes in a water matrix.  The MDLs for the components of a  specific sample may
differ  from those listed in  Table  1  because  MDLs  depend  on  the  nature  of
interferences in the sample  matrix.  Table  2  lists  the estimated quantitation
limits (EQL) for other matrices.

      1.3   When this method  is  used to  analyze for any  or  all  of the target
analytes, compound identification should be supported by at least one additional
qualitative technique.  This  method  describes  conditions for parallel column,
dual electron capture detector analysis which  fulfills the  above requirement.
Retention time  information obtained  on two  megabore  fused-silica open tubular
columns is given in Table  1.  Alternatively, gas chromatography/mass spectrometry
could be used for compound confirmation.

      1.4   The  following compounds,  bis(2-n-butoxyethyl)   phthalate,  bis(2-
ethoxyethyl)  phthalate, bis(2-methoxyethyl)  phthalate, bis(4-tnethyl-2-pentyl)
phthalate,   diamyl  phthalate,   dicyclohexyl   phthalate,   dihexyl   phthalate,
diisobutyl  phthalate,  dinonyl  phthalate,  and hexyl  2-ethylhexyl  phthalate can
also be analyzed by this method and may be used as surrogates.

      1.5   This method  is  restricted  to use  by  or under  the  supervision  of
analysts experienced  in  the  use  of gas chromatographs and  skilled in  the
interpretation of gas  chromatograms.  Each analyst must demonstrate the ability
to generate acceptable results with this  method.
                                   8061 - 1                         Revision 0
                                                                September 1994

-------
2.0   SUMMARY OF METHOD

      2.1   A measured volume  or  weight of sample  (approximately  1  liter for
liquids,  10  to  30  grams  for  solids  and  sludges)  is  extracted by  using the
appropriate sample extraction technique specified in Methods 3510,  3540, 3541,
and 3550.  Method 3520 is  not recommended for the extraction of aqueous samples
because the longer chain esters  (dihexyl  phthalate, bis(2-ethylhexyl) phthalate,
di-n-octyl phthalate,  and  dinonyl phthalate) tend to  adsorb to the glassware and
consequently, their extraction  recoveries are <40 percent.  Aqueous samples are
extracted at a  pH of  5 to 7,  with methylene chloride,  in a separatory funnel
(Method 3510).  Alternatively, particulate-free aqueous samples could be filtered
through membrane disks that contain C18-bonded  silica.  The phthalate esters are
retained by the silica and,  later eluted with  acetonitrile.  Solid samples are
extracted with  hexane/acetone  (1:1)  or methylene chloride/acetone  (1:1)  in a
Soxhlet extractor  (Methods 3540/3541) or with  an  ultrasonic extractor (Method
3550).   After  cleanup,  the  extract  is analyzed by  gas  chromatography  with
electron capture detection (GC/ECD).

      2.2   The  sensitivity  of Method  8061  usually depends on the  level  of
interferences rather than  on instrumental limitations.  If  interferences prevent
detection of the analytes, cleanup  of  the sample extracts  is necessary.  Either
Method 3610 or 3620 alone or followed by Method 3660,  Sulfur Cleanup, may be used
to eliminate interferences in the analysis.  Method 3640,  Gel Permeation Cleanup,
is applicable for samples that  contain high amounts  of lipids and waxes.


3.0   INTERFERENCES

      3.1   Refer to Methods  3500,  3600, and 8000.

      3.2   Interferences coextracted from the  samples  will  vary considerably
from waste to waste.  While general cleanup techniques are referenced or provided
as part of this  method, unique samples  may require additional cleanup approaches
to achieve desired sensitivities for the target analytes.

      3.3   Glassware  must  be scrupulously  clean.    All  glassware  require
treatment in a muffle  furnace at 400 "C for  2 to  4 hrs, or  thorough rinsing with
pesticide-grade  solvent,  prior to  use.   Refer  to  Chapter 4,  Sec.  4.1.4, for
further details  regarding  the cleaning  of glassware.  Volumetric glassware should
not be heated in a muffle furnace.

      If Soxhlet extractors  are baked  in the muffle  furnace, care must be taken
to ensure that  they are  dry  (breakage may  result if any  water  is  left  in the
side-arm).  Thorough rinsing with hot tap water,  followed by deionized water and
acetone  is  not  an adequate decontamination procedure.    Even  after  a Soxhlet
extractor was refluxed with acetone for  three days,  with daily solvent changes,
the concentrations  of bis(2-ethylhexyl)  phthalate  were as high  as 500 ng per
washing.  Storage of glassware  in the  laboratory introduces contamination, even
if the glassware is wrapped in  aluminum  foil.  Therefore,  any glassware used in
Method 8061 should be cleaned  immediately prior to use.

      3.4   Florisil and alumina may be  contaminated with  phthalate esters and,
therefore,  use   of these materials  in  sample  cleanup   should   be  employed


                                   8061 - 2                         Revision 0
                                                                September 1994

-------
cautiously.  If these materials are used,  they must be obtained packaged  in glass
(plastic  packaging  will  contribute  to  contamination with  phthalate esters).
Washing of these  materials  prior to use with  the  solvent(s)  used for elution
during extract cleanup was found helpful,  however, heating at 320 °C for Florisil
and  210  °C for  alumina  is  recommended.    Phthalate  esters were  detected in
Florisil  cartridge method blanks  at  concentrations  ranging  from 10 to 460 ng,
with 5 phthalate  esters  in  the 105 to 460 ng  range.  Complete removal  of the
phthalate esters  from  Florisil cartridges  does  not seem possible,  and  it is
therefore  desirable  to  keep  the  steps  involved in  sample preparation  to  a
minimum.

      3.5   Paper thimbles  and filter paper must be  exhaustively washed with the
solvent that  will  be used in the sample extraction.   Soxhlet  extraction of paper
thimbles and  filter paper for  12 hrs with fresh solvent  should be repeated for
a minimum of three times.   Method  blanks  should  be  obtained before any of the
precleaned thimbles or filter papers are used.   Storage  of precleaned thimbles
and  filter paper  in   precleaned   glass  jars   covered  with  aluminum foil  is
recommended.

      3.6   Glass wool  used  in  any step  of   sample  preparation  should  be  a
specially treated pyrex wool,  pesticide  grade, and  must  be  baked  at 400°C for
4 hrs. immediately prior to use.

      3.7   Sodium sulfate must be  obtained packaged in glass (plastic packaging
will contribute to contamination with phthalate esters),  and  must be purified by
heating at 400 °C  for 4 hrs. in a shallow tray,  or by precleaning with methylene
chloride (Sec. 5.3).  To  avoid recontamination, the precleaned material must be
stored in  glass-stoppered glass bottles, or glass bottles  covered with precleaned
aluminum foil.  The storage period should not exceed two weeks.   To minimize
contamination, extracts should be dried directly  in the glassware  in which they
are  collected  by  adding  small  amounts  of  precleaned sodium  sulfate  until  an
excess of free flowing  material is noted.

      3.8   The presence of elemental sulfur will  result  in large peaks which
often mask the region  of the  compounds  eluting  before  dicyclohexyl  phthalate
(Compound No.  14) in the gas  chromatograms  shown in  Figure  1.   Method  3660 is
suggested for removal  of sulfur.

      3.9   Waxes and  lipids  can  be removed by  Gel  Permeation Chromatography
(Method 3640).  Extracts containing high concentrations  of lipids  are viscous,
and may even  solidify  at room temperature.


4.0   APPARATUS AND MATERIALS

      4.1   Gas chromatography

            4.1.1 Gas   chromatograph,  analytical  system  complete  with  gas
      chromatograph suitable  for on-column  and split/splitless injections and
      all   required  accessories,   including   detector,   analytical   columns,
      recorder, gases,  and  syringes.  A data system for  measuring  peak heights
      and/or  peak areas is  recommended.
                                   8061 - 3                         Revision 0
                                                                September 1994

-------
                  4.1.1.1     Eight inch injection tee (Supelco, Inc., Catalog
            No. 2-3665, or equivalent)  or glass  Y  splitter for megabore columns
            (J&W Scientific,-"press-fit", Catalog  No.  705-0733, or equivalent).

            4.1.2 Columns

                  4.1.2.1     Column 1, 30  m x 0.53 mm ID, 5% phenyl/95% methyl
            silicone fused-silica open  tubular column  (DB-5, J&W Scientific, or
            equivalent), 1.5 p,m film thickness.

                  4.1.2.2     Column 2,  30  m  x  0.53  mm ID,  14%  cyanopropyl
            phenyl  silicone  fused-silica  open  tubular  column  (DB-1701,  J&W
            Scientific, or equivalent), 1.0 fj,m  film thickness.

            4.1.3 Detector - Dual electron  capture detector (ECD)

      4.2   Glassware, see Methods 3510, 3540, 3541,  3550, 3610, 3620, 3640, and
3660 for specifications.

      4.3   Kuderna-Danish (K-D) apparatus.

            4.3.1 Concentrator tube - 10 ml graduated  (Kontes K-570050-1025 or
      equivalent).  A  ground  glass stopper  is  used to prevent  evaporation of
      extracts.

            4.3.2 Evaporation flask  -  500  mL (Kontes  K-570001-500  or equiva-
      lent).  Attach to concentrator tube with springs, clamps, or equivalent.

            4.3.3 Snyder column  - Three ball macro  (Kontes  K-503000-0121 or
      equivalent).

            4.3.4 Snyder  column  -  Two  ball  micro  (Kontes  K-569001-0219 or
      equivalent).

            4.3.5 Springs - 1/2 inch (Kontes K-662750 or equivalent).

      4.4   Boiling chips,  approximately 10/40 mesh.  Heat to  400  °C for 30 min,
or Soxhlet-extract with methylene chloride  prior to use.

      4.5   Water  bath,  heated,  with  concentric  ring  cover,  capable  of
temperature control (+ 2°C).


5.0   REAGENTS

      5.1   Reagent grade chemicals shall be used in all tests.  Unless otherwise
indicated, it is intended that  all  reagents  shall  conform to the specifications
of the Committee on Analytical  Reagents  of the American Chemical Society, where
such specifications are  available.   Other  grades  may  be used,  provided it is
first ascertained that the  reagent is of sufficiently  high purity to permit its
use without lessening the accuracy of the determination.
                                   8061 - 4                         Revision 0
                                                                September 1994

-------
      5.2   Organic-free reagent water.   All  references  to water  in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.3   Sodium sulfate (granular, anhydrous),  Na2S04.  Purify by heating at
400 °C for 4 hours in a  shallow  tray, or  by precleaning  the sodium sulfate with
methylene chloride.   If the sodium sulfate is precleaned with methylene chloride,
a method blank must be analyzed,  demonstrating that there is no interference from
the sodium sulfate.

      5.4   Solvents:

            5.4.1 Hexane, C6H14  -  Pesticide quality, or  equivalent.

            5.4.2 Methylene chloride, CH2C12 - Pesticide  quality, or equivalent.

            5.4.3 Acetone, CH3COCH3  - Pesticide  quality,  or equivalent.

            5.4.4 Acetonitrile,  CH3CN -  HPLC  grade.

            5.4.5 Methanol, CH3OH -  HPLC  grade.

            5.4.6 Diethyl Ether,  C2H5OC2H5  -  Pesticide quality,  or equivalent.
      Must  be  free of  peroxides,  as indicated by  test strips  (EM  Quant,  or
      equivalent).   Procedures  for  removal of  peroxides are  provided with the
      test strips.  After cleanup, 20 ml of ethyl  alcohol preservative must be
      added to each liter of ether.

      5.5   Stock standard solutions:

            5.5.1 Prepare  stock  standard  solutions  at  a  concentration  of
      1000 mg/L by dissolving 0.0100 g of assayed reference material in hexane,
      and diluting to volume  in  a  10 ml volumetric flask.  When compound purity
      is assayed  to  be  96 percent or greater,  the weight can  be used without
      correction  to  calculate  the  concentration  of   the   stock  standard.
      Commercially  prepared   stock   standard  solutions   can   be   used  at  any
      concentration  if   they are  certified  by  the  manufacturer  or  by  an
      independent source.

            5.5.2 Transfer the  stock standard  solutions  into  glass  vials  with
      Teflon lined screw-caps or  crimp  tops.   Store at 4 °C  and protect  from
      light.  Stock  standard solutions  should be checked periodically  by gas
      chromatography for  signs  of degradation  or  evaporation,  especially  just
      prior to preparation of calibration standards.

            5.5.3 Stock  standard solutions  must be replaced  after 6 months,  or
      sooner if comparison with check standards indicates a  problem.

      5.6   Calibration   standards:  Calibration standards  are  prepared   at  a
minimum of five concentrations for each  parameter  of interest through dilution
of the stock standard solutions with hexane.   One  of the concentrations should
be at a concentration near, but above, the method detection limit.  The remaining
concentrations should correspond to the  expected range of concentrations  found
in real  samples,  or  should define  the working  range  of the GC.   Calibration


                                   8061  - 5                         Revision 0
                                                                September  1994

-------
solutions must be  replaced  after  1  to 2 months, or  sooner  if comparison with
calibration verification standards indicates a problem.

      5.7   Internal standards (if internal standard calibration is used):  To
use this approach, the analyst must select one or more internal standards that
are similar in analytical behavior to  the  compounds  of interest.   The analyst
must further demonstrate that the measurement  of the internal  standard is not
affected by method or matrix interferences.  Benzyl  benzoate  has  been tested and
found appropriate for Method 8061.

            5.7.1 Prepare a spiking  solution  of benzyl benzoate  in  hexane at
      5000 mg/L.   Addition of 10 juL of this solution to 1  ml of sample extract
      is recommended.  The spiking concentration of  the  internal  standard should
      be kept constant  for  all  samples and calibration standards.   Store the
      internal standard  spiking  solution  at 4  °C  in glass vials  with Teflon
      lined screw-caps  or crimp  tops.   Standard solutions  should  be replaced
      when ongoing QC (Sec.  8) indicates a problem.

      5.8   Surrogate standards:  The analyst should monitor the performance of
the extraction, cleanup  (when used), analytical  system,  and the effectiveness of
the method in  dealing  with each sample  matrix by  spiking each sample,  standard,
and blank with surrogate compounds. Three surrogates may be  used for Method 8061
in  addition  to   those   listed  in  Sec.   1.4:  diphenyl  phthalate,   diphenyl
isophthalate,  and dibenzyl phthalate.  However,  the  compounds listed in Sec. 1.4
are recommended.

            5.8.1 Prepare a surrogate  standard spiking solution,  in acetone,
      which contains  50 ng/juL  of each compound.   Addition  of  500 /iL  of this
      solution to 1 L  of water or  30 g  solid sample is equivalent to 25 ^tg/L of
      water or 830  M9A9 °f solid sample.   The spiking concentration  of the
      surrogate standards may be  adjusted  accordingly,  if  the final  volume of
      extract   is  reduced below  2 ml  for  water  samples  or  10 ml  for  solid
      samples. Store  the surrogate spiking solution at  4 °C  in glass vials with
      Teflon lined  screw-caps  or  crimp tops.   The solution must  be replaced
      after 6  months,  or sooner if ongoing QC (Sec. 8)  indicates problems.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See the introductory  material  to this chapter, Organic  Analytes,
Sec. 4.1.
7.0   PROCEDURE

      7.1   Extraction:

            7.1.1 Refer to Chapter Two  for guidance on choosing the appropriate
      extraction procedure.  In general, water  samples are extracted at a pH of
      5 to  7  with methylene  chloride  in  a  separatory  funnel  (Method  3510).
      Method 3520  is not  recommended  for the  extraction  of  aqueous  samples
      because  the  longer  chain  esters  (dihexyl  phthalate  bis(2-ethylhexyl)
      phthalate, di-n-octyl phthalate,  and dinonyl phthalate) tend to adsorb to


                                   8061 -  6                          Revision 0
                                                                September 1994

-------
the  glassware   and   consequently,   their   extraction   recoveries  are
<40 percent.   Solid  samples are extracted  with  hexane/acetone (1:1) or
methylene  chloride/acetone   (1:1)   in   a   Soxhlet  extractor  (Methods
3540/3541) or  with  an ultrasonic extractor  (Method  3550).   Immediately
prior  to extraction,  spike 500 p,l  of  the surrogate  standard  spiking
solution (concentration = 50 ng//iL)  into 1  L aqueous  sample or 30 g solid
sample.

      7.1.2 Extraction   of  particulate-free   aqueous    samples   using
C18-extraction  disks  (optional):

            7.1.2.1     Disk preconditioning: Place the C18-extraction disk
      into the filtration apparatus  and  prewash the disk with 10 to 20 ml
      of acetonitrile.  Apply vacuum to pull the solvent  through the disk.
      Maintain vacuum to pull air through for 5 min.  Follow with 10 ml of
      methanol.  Apply  vacuum  and pull  most of  the methanol  through the
      disk.  Release vacuum before the disk gets dry.  Follow with 10 ml
      organic-free reagent water.  Apply vacuum and pull  most of the water
      through the disk.  Release the vacuum before the disk gets dry.

            7.1.2.2     Sample preconcentration: Add 2.5 ml of methanol to
      the  500  ml  aqueous sample  in  order  to get  reproducible results.
      Pour the  sample  into the filtration  apparatus.   Adjust  vacuum so
      that  it  takes  approximately 20 min to process  the  entire sample.
      After all of the sample has passed through the membrane disk, pull
      air through the disk for 5  to  10 min.  to remove any residual  water.

            7.1.2.3     Sample elution:  Break the vacuum and place the tip
      of the filter base into the test tube that is contained inside the
      suction flask.  Add 10 ml of acetonitrile to the graduated funnel,
      making sure  to  rinse the walls  of the graduated funnel  with  the
      solvent.   Apply  vacuum to pass  the acetonitrile  through the membrane
      disk.

            7.1.2.4     Extract concentration (if necessary): Concentrate
      the extract to  2  ml  or less,  using either  the  micro Snyder  column
      technique  (Sec.  7.1.2.4.1)  or  nitrogen  blowdown technique  (Sec.
      7.1.2.4.2).

                  7.1.2.4.1   Micro  Snyder Column Technique

                        7.1.2.4.1.1  Add one or two  clean boiling chips to
                  the concentrator tube  and attach a two  ball micro Snyder
                  column.   Prewet the  column by adding about  0.5  ml of
                  acetonitrile to the top  of the  column.   Place the K-D
                  apparatus in a hot water bath (15-20°C  above the boiling
                  point of the solvent) so that the concentrator tube is
                  partially immersed in  the hot water and the entire lower
                  rounded surface of the flask is bathed with hot  vapor.
                  Adjust the vertical position of  the  apparatus  and  the
                  water  temperature,   as   required,   to   complete  the
                  concentration in 5-10 minutes.   At the  proper rate of
                  distillation  the  balls  of the  column  will  actively


                             8061 -  7                         Revision 0
                                                          September 1994

-------
                        chatter,  but  the  chambers will  not  flood.    When  the
                        apparent  volume of liquid reaches 0.5  mL, remove the K-D
                        apparatus from the water  bath and allow it to drain and
                        cool  for  at least 10 minutes.  Remove the Snyder column
                        and rinse  the flask and  its  lower  joints  with  about
                        0.2 ml of  solvent  and add  to  the concentrator  tube.
                        Adjust the final  volume to 1.0-2.0 ml with solvent.

                        7.1.2.4.2   Nitrogen Slowdown  Technique

                              7.1.2.4.2.1  Place the  concentrator tube  in a warm
                        water  bath  (approximately  35 °C)  and  evaporate  the
                        solvent volume to  the required  level  using  a  gentle
                        stream of clean, dry nitrogen (filtered through a column
                        of activated carbon).

                              CAUTION:    Do not  use plasticized tubing between
                                          the carbon trap and the  sample.

                              7.1.2.4.2.2  The internal  wall of the tube must be
                        rinsed down several  times with  acetonitrile  during the
                        operation.   During evaporation,  the solvent level in the
                        tube must be positioned to prevent water from condensing
                        into the  sample (i.e., the solvent level should be below
                        the level of  the water bath).   Under normal operating
                        conditions, the extract should  not be allowed to become
                        dry.

      7.2   Solvent Exchange:  Prior to Florisil  cleanup or gas  chromatographic
analysis, the methylene chloride and methylene chloride/acetone extracts obtained
in Sec.  7.1.1  must be exchanged to hexane,  as described in Sees.  7.2.1 through
7.2.3.   Exchange  is not  required   for the  acetonitrile extracts obtained  in
Sec. 7.1.2.4.

            7.2.1  Add one or two  clean boiling chips to the flask  and attach a
      three ball Snyder column.  Concentrate the  extract  as  described in  Sec.
      7.1.2.4.1, using  1  ml  of methylene chloride  to  prewet the column,  and
      completing the concentration in 10-20  minutes.   When the  apparent volume
      of liquid reaches  1-2 ml, remove the  K-D apparatus from  the water bath and
      allow it to drain and cool  for at  least 10  minutes.

            7.2.2 Momentarily remove the  Snyder column, add 50  ml  of hexane, a
      new boiling chip,  and  attach the  macro Snyder column.   Concentrate  the
      extract  as described in Sec. 7.1.2.4.1, using 1 ml of hexane to prewet the
      Snyder column,  raising the temperature of the water bath, if necessary,  to
      maintain proper distillation, and completing the  concentration  in  10-20
      minutes.  When the  apparent volume of liquid  reaches 1-2 ml,  remove  the
      K-D apparatus and allow it  to drain  and cool for  at least 10 min.

            7.2.3 Remove  the Snyder  column  and  rinse  the flask and  its  lower
      joint into the  concentrator tube with 1 to 2 mL hexane.  A 5 ml syringe is
      recommended for this operation.  Adjust  the extract volume to  2 mL  for
      water samples,  using  either the micro  Snyder  column technique  (Sec.


                                   8061  -  8                          Revision 0
                                                                September  1994

-------
7.1.2.4.1) or nitrogen blowdown technique {Sec. 7.1.2.4.2), or 10 ml for
solid samples.   Stopper the concentrator tube and store at 4 °C if further
processing will be performed immediately.  If the extract will be stored
for two days or longer,  it  should  be  transferred  to a glass vial  with a
Teflon lined screw-cap or crimp top.  Proceed  with  the gas chromatographic
analysis.

7.3   Cleanup/Fractionation:

      7.3.1 Cleanup may  not be necessary  for extracts  from a relatively
clean  sample   matrix.      If   polychlorinated   biphenyls   (PCBs)   and
organochlorine pesticides are known to be present in the sample,  use the
procedure outlined in Methods 3610 or  3620.   When using column  cleanup,
collect  Fraction   1  by  eluting  with  140 ml  (Method  3610)  or  100 ml
(Method 3620) of  20-percent diethyl  ether in hexane.   Note that,  under
these  conditions,   bis(2-methoxyethyl)   phthalate,  bis(2-ethoxyethyl)
phthalate, and bis(2-n-butoxyethyl) phthalate are not recovered  from the
Florisil  column.  The elution patterns and compound recoveries are given
in Table 3.

      7.3.2 Methods 3610 and  3620  also describe   procedures  for  sample
cleanup  using   Alumina   and  Florisil  Cartridges.   With   this  method,
bis(2-methoxyethyl)   phthalate,    bis(2-ethoxyethyl)   phthalate,    and
bis(2-n-butoxyethyl)  phthalate are recovered quantitatively.

7.4   Gas chromatographic conditions (recommended):

      7.4.1 Column 1  and Column 2 (Sec. 4.1.2):

      Carrier gas (He) =            6 inL/min.
      Injector temperature  =        250 °C.
      Detector temperature  =        320 °C.
      Column temperature:
            Initial temperature =   150 °C,  hold for 0.5 min.
            Temperature program =   150  °C   to 220  °C  at  5  °C/min.,
                                    followed by  220 °C  to  275  °C  at  3
                                    °C/min.
            Final  temperature =     275 °C hold for 13 min.

      7.4.2 Table  1  gives  the  retention  times  and  MDLs  that  can  be
achieved by this  method  for the  16  phthalate esters.   An example  of the
separations achieved  with the DB-5 and DB-1701 fused-silica open  tubular
columns is shown  in Figure  1.

7.5   Calibration:

      7.5.1 Refer to  Method 8000 for proper calibration techniques.   Use
Tables  1   and  2  for guidance  on  selecting the  lowest  point  on  the
calibration curve.

      7.5.2 The procedure  for internal  or  external calibration may  be
used.   Refer  to  Method 8000  for  the  description  of  each  of  these
procedures.


                             8061 - 9                         Revision 0
                                                          September 1994

-------
      7.6   Gas chromatographic analysis:

            7.6.1 Refer to Method 8000.   If  the  internal  standard calibration
      technique is used, add 10 //L  of  internal  standard solution at 5000 mg/L
      to the sample prior to injection.

            7.6.2 Follow Method 8000 for  instructions on the analysis sequence,
      appropriate dilutions,  establishing daily  retention  time  windows,  and
      identification criteria.

            7.6.3 Record the  sample volume  injected  and  the  resulting  peak
      areas.

            7.6.4 Using  either  the  internal   or  the  external  calibration
      procedure (Method 8000),  determine the identity and the quantity of each
      component  peak  in  the  sample  chromatogram  which corresponds  to  the
      compounds used for calibration purposes.

            7.6.5 If the response of a peak exceeds the working  range  of the
      system, dilute the extract and reanalyze.

            7.6.6 Identify compounds in  the  sample  by  comparing the retention
      times of the peaks in the sample chromatogram with those of the peaks in
      standard  chromatograms.     The  retention  time  window  used  to  make
      identifications  is  based  upon  measurements  of  actual   retention  time
      variations over the course of  10  consecutive injections.  Three times the
      standard deviation  of the  retention time  can  be used  to  calculate  a
      suggested window size.


8.0   QUALITY CONTROL

      8.1   Refer  to  Chapter  One  for specific  quality control  procedures.
Quality control to validate sample extraction is covered in  Method 3500 and in
the extraction method utilized.  If extract cleanup was  performed,  follow the QC
specified in Method 3600 and in the specific cleanup method.

      8.2   Quality control  required to evaluate the GC system operation is found
in Method 8000.

            8.2.1 The  quality  control  check sample concentrate  (Method  8000)
      should contain the test compounds at 5 to  10 ng/^L.

      8.3   Calculate the recoveries  of the surrogate compounds for all samples,
method blanks, and method spikes.  Determine if the recoveries are within limits
established by performing QC procedures outlined in  Method  8000.

            8.3.1 If the recoveries  are  not within limits,   the  following are
      required:

                  8.3.1.1     Make sure  there are  no errors  in calculations,
            surrogate solutions and internal  standards.  Also check instrument
            performance.


                                  8061  -  10                         Revision 0
                                                                September 1994

-------
                  8.3.1.2     Recalculate the data and/or reanalyze the extract
            if any of the above checks reveal a problem.

                  8.3.1.3     Reextract and reanalyze the sample  if none of the
            above are a problem,  or flag  the  data  as  "estimated concentration."

      8.4   An  internal  standard  peak area  check must  be performed  on  all
samples.  The internal  standard  must be evaluated  for  acceptance by determining
whether the measured  area  for the  internal  standard  deviates  by  more than 30
percent from  the average  area  for  the  internal  standard  in  the calibration
standards.  When the internal standard  peak area is outside  that  limit,  all
samples that fall outside the QC criteria must be reanalyzed.

      8.5   GC/MS confirmation:  Any compounds confirmed  by two  columns may also
be confirmed by GC/MS if the concentration  is sufficient  for detection by GC/MS
as determined by the laboratory-generated detection limits.

            8.5.1 The GC/MS would normally require a minimum concentration of 10
      ng//nL in the final extract for each single-component compound.

            8.5.2 The sample extract and  associated blank should be analyzed by
      GC/MS as per Sec.  7.0  of Method  8270.   Normally,  analysis  of  a blank is
      not required for confirmation analysis, however, analysis for phthalates
      is  a  special  case because of the  possibility  for sample  contamination
      through septum punctures,  etc.

            8.5.3 A reference standard  of the compound must  also be analyzed by
      GC/MS.   The  concentration  of  the  reference  standard  must   be  at  a
      concentration that would demonstrate  the ability to confirm the phthalate
      esters identified by GC/ECD.

      8.6   Include a mid-concentration calibration standard after  each group of
20 samples  in   the   analysis  sequence.     The   response   factors   for   the
mid-concentration calibration must  be within  + 15  percent of the average values
for the multiconcentration calibration.

      8.7   Demonstrate  through  the analyses of  standards that  the  Florisil
fractionation scheme  is reproducible. When using the fractionation  schemes given
in Methods 3610  or 3620,  batch-to-batch  variations in the  composition  of  the
alumina or  Florisil  material may  cause  variations  in  the recoveries  of  the
phthalate esters.


9.0   METHOD PERFORMANCE

      9.1   The MDL is defined in Chapter One. The MDL concentrations listed in
Table 1 were  obtained  using  organic-free  reagent water.   Details  on  how  to
determine MDLs are given in Chapter One.   The MDL actually achieved  in a given
analysis  will  vary,  as  it  is dependent  on  instrument sensitivity  and  matrix
effects.

      9.2   This method  has been tested  in a single laboratory by using different
types of  aqueous samples and  solid  samples which  were fortified  with  the  test


                                   8061 -  11                         Revision 0
                                                                September  1994

-------
compounds at two concentrations.  Single-operator precision, overall precision,
and method  accuracy  were  found to  be  related  to  the  concentration  of the
compounds and  the type  of  matrix.    Results  of the  single-laboratory method
evaluation are presented in Tables 4 and 5.

      9.3   The  accuracy  and  precision  obtained  is  determined by  the sample
matrix,  sample  preparation  technique,  cleanup  techniques,   and  calibration
procedures used.


10.0  REFERENCES

1.    Glazer,  J.A.;  Foerst, G.D.;  McKee, G.D.; Quave, S.A.,  and  Budde, W.L.,
      "Trace Analyses  for  Wastewaters," Environ.  Sci. and  Techno!.  15: 1426,
      1981.

2.    Lopez-Avila, V.,  Baldin,  E.,  Benedicto,  J.,  Milanes, J.,  and  Beckert,
      W.F.,  "Application of Open-Tubular Columns to SW-846 GC Methods", EMSL-Las
      Vegas, 1990.

3.    Beckert, W.F. and Lopez-Avila, V., "Evaluation of SW-846 Method 8060 for
      Phthalate  Esters",   Proceedings   of  Fifth  Annual   Testing  and  Quality
      Assurance Symposium, USEPA, 1989.
                                   8061  -  12                         Revision 0
                                                                September 1994

-------
                                        TABLE  1.
GAS CHROMATOGRAPHIC RETENTION TIMES AND METHOD DETECTION LIMITS FOR THE PHTHALATE ESTERS8
Compound
No.
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
IS
SU-1
SU-2
SU-3
Compound name
Dimethyl phthalate
Diethyl phthalate
Diisobutyl phthalate
Di-n-butyl phthalate
Bis(4-methyl-2-pentyl) phthalate
Bis(2-methoxyethyl) phthalate
Diamyl phthalate
Bis(2-ethoxyethyl) phthalate
Hexyl 2-ethylhexyl phthalate
Dihexyl phthalate
Butyl benzyl phthalate
Bis(2-n-butoxyethyl ) phthalate
Bis(2-ethylhexyl) phthalate
Dicyclohexyl phthalate
Di-n-octyl phthalate
Dinonyl phthalate
Benzyl benzoate
Diphenyl phthalate
Diphenyl isophthalate
Dibenzyl phthalate
Chemical
Abstract
Registry
No.
131-11-3
84-66-2
84-69-5
84-74-2
146-50-9
117-82-8
131-18-0
605-54-9
75673-16-4
84-75-3
85-68-7
117-83-9
117-81-7
84-61-7
117-84-0
84-76-4
120-51-4
84-62-8
744-45-6
523-31-9
Retention time8
(min)
Column 1
7.06
9.30
14.44
16.26
18.77
17.02
20.25
19.43
21.07
24.57
24.86
27.56
29.23
28.88
33.33
38.80
12.71
29.46
32.99
34.40
Column 2
6.37
8.45
12.91
14.66
16.27
16.41
18.08
18.21
18.97
21.85
23.08
25.24
25.67
26.35
29.83
33.84
11.07
28.32
31.37
32.65
MDLb
Liquid
(ng/L)
640
250
120
330
370
510
110
270
130
68
42
84
270
22
49
22
c
c
c
c
                                        8061 - 13
    Revision 0
September 1994

-------
                                     Table 1. (continued)

Column 1 is a 30 m x 0.53 mm ID DB-5 fused-silica open tubular column (1.5 //m film thickness).
Column 2 is a 30 m 0.53 mm ID DB-1701 fused-silica open tubular column (1.0 urn film thickness).
Temperature program is 150°C (0.5 min hold) to 220°C at 5°C/min, then to 275'C (13 min hold) at
3°C/min.  An 8-in Supelco injection tee or a J&W Scientific press fit glass inlet splitter is used
to connect the two columns to the injection port of a gas chromatograph.  Carrier gas helium at
6 mL/min; makeup gas nitrogen at 20 mL/min; injector temperature 250°C; detector temperature
320°C.

MDL is the method detection limit.  The MDL was determined from the analysis of seven replicate
aliquots of organic-free reagent water processed through the entire analytical method (extraction,
Florisil cartridge cleanup, and GC/ECD analysis using the single column approach:  DB-5 fused-
silica capillary column).  MDL = t(lvl 099) x SD where t(n.., 0991  is  the  student's  t value  appropriate
for a 99 percent confidence interval and a standard deviation with n-1 degrees of freedom, and SD
is the standard deviation of the seven replicate measurements.  Values measured were not corrected
for method blanks.

Not applicable.
                                           8061  -  14                                       Revision 0
                                                                                       September 1994

-------
                             TABLE 2.
     ESTIMATED QUANTITATION  LIMITS (EQL)  FOR VARIOUS MATRICES6
Matrix                                                Factor
Groundwater                                                10
Low-concentration soil by ultrasonic extraction           670
  with GPC cleanup
High-concentration soil  and sludges by ultrasonic      10,000
  extraction
Non-water miscible waste                              100,000
EQL = [Method detection limit (see Table 1)] X [Factor found in this
table].  For non-aqueous samples, the factor is on a wet-weight basis.
Sample EQLs are highly matrix-dependent.  The EQLs determined herein are
provided for guidance and may not always be achievable.
                             8061  -  15                         Revision 0
                                                          September 1994

-------
                                   TABLE 3.
   AVERAGE RECOVERIES OF METHOD 8061 COMPOUNDS  USING METHODS  3610  AND  3620
Compound
Dimethyl phthalate
Diethyl phthalate
Diisobutyl phthalate
Di-n-butyl phthalate
Bis(4-methyl-2-pentyl) phthalate
Bis(2-methoxyethyl) phthalate
Diamyl phthalate
Bis(2-ethoxyethyl) phthalate
Hexyl 2-ethylhexyl phthalate
Dihexyl phthalate
Benzyl butyl phthalate
Bis(2-n-butoxyethyl) phthalate
Bis(2-ethylhexyl) phthalate
Dicyclohexyl phthalate
Di-n-octyl phthalate
Dinonyl phthalate
Alumina
col umn8
64.5
62.5
77.0
76.5
89.5
70.5
75.0
67.0
90.5
73.0
87.0
62.5
91.0
84.5
108
71.0
Florisil
column"
40.0
57.0
80.0
85.0
84.5
0
81.5
0
105
74.5
90.0
0
82.0
83.5
115
72.5
Alumina
cartridge6
101
103
104
108
103
64.1°
103
111
101
108
103
108
97.6
97.5
112
97.3
Florisil
cartridge6
89.4
97.3
91.8
102
105
78. 3e
94.5
93.6
96.0
96.8
98.6
91.5
97.5
90.5
97.1
105
a 2 determinations; alumina and Florisil chromatography performed according
  to Methods 3610 and 3620, respectively.

b 2 determinations, using 1 g alumina cartridges; Fraction 1 was eluted with
  5 ml of 20-percent acetone in hexane.  40 jug of each component was spiked
  per cartridge.

c 36.8 percent was recovered by elution with an additional 5 ml of
  20-percent acetone in hexane.

d 2 determinations, using 1 g Florisil cartridges; Fraction 1 was eluted
  with 5 ml of 10-percent acetone in hexane.  40 /^g of each component was
  spiked per cartridge.

e 14.4 percent was recovered by elution with an additional 5 ml of
  10-percent acetone in hexane.
                                   8061  -  16
    Revision 0
September 1994

-------
                                                     TABLE  4.
                            ACCURACY  AND  PRECISION  DATA FOR  METHOD  3510  AND  METHOD  8061£
Spike Concentration
(20 uq/U
Estuarine
Compound
Dimethyl phthalate
Diethyl phthalate
Diisobutyl phthalate
Di-n-butyl phthalate
Bis(4-methyl-2-pentyl) phthalate
Bis(2-methoxyethyl) phthalate
Diamyl phthalate
Bis(2-ethoxyethyl) phthalate
Hexyl 2-ethylhexyl phthalate
Dihexyl phthalate
Benzyl butyl phthalate
Bis(2-n-butoxyethyl) phthalate
Bis(2-ethylhexyl) phthalate
Dicyclohexyl phthalate
Di-n-octyl phthalate
Dinonyl phthalate
Surrogates:
Diphenyl phthalate
Diphenyl isophthalate
Dibenzyl phthalate
water
84.0
71.2
76.0
83.2
78.6
73.8
78.2
75.6
84.7
79.8
84.1
78.5
81.4
77.4
74.9
59.5

98.5
95.8
93.9
(4.1)
(3.8)
(6.5)
(6.5)
(2.6)
(1.0)
(7.3)
(3.3)
(5.3)
(7.2)
(6.4)
(3.5)
(4.1)
(6.5)
(4.9)
(6.1)

(2.6)
(1.9)
(4.4)
Leachate
98.9
82.8
95.3
97.5
87.3
87.2
92.1
90.8
91.1
102
105
92.3
93.0
88.2
87.5
77.3

113
112
112
(19.6)
(19.3)
(16.9)
(22.3)
(18.2)
(21.7)
(21.5)
(22.4)
(27.5)
(21.5)
(20.5)
(16.1)
(15.0)
(13.2)
(18.7)
(4.2)

(14.9)
(11-7)
(14.0)
Estuarine
Groundwater
87.1
88.5
92.7
91.0
92.6
82.4
88.8
86.4
81.4
90.9
89.6
89.3
90.5
91.7
87.2
67.2

110
109
106
(8.1)
(15.3)
(17.1)
(10-7)
(13.7)
(4.4)
(7.5)
(5.8)
(17.6)
(7.6)
(6.1)
(3.6)
(4.9)
(15.2)
(3.7)
(8.0)

(3.3)
(3.3)
(3.8)
Sp
water
87.1
71.0
99.1
87.0
97.4
82.5
89.2
88.7
107
90.1
92.7
86.1
86.5
87.7
85.1
97.2

110
104
111
(7.5)
(7.7)
(19.0)
(8.0)
(15.0)
(5.5)
(2.8)
(4.9)
(16.8)
(2.4)
(5.6)
(6.2)
(6.9)
(9.6)
(8.3)
(7.0)

(12.4)
(5.9)
(5.9)
ike Concentration
(60 uq/L)
Leachate
112
88.5
100
106
107
99.0
112
109
117
109
117
107
108
102
105
108

95.1
97.1
93.3
(17-5)
(17.9)
(9.6)
(17.4)
(13.3)
(13.7)
(14.2)
(14.6)
(11.4)
(20.7)
(24.7)
(15.3)
(15.1)
(14.3)
(17.7)
(17.9)

(7.2)
(7.1)
(9.5)
Groundwater
90.9 (4.5)
75.3 (3.5)
83.2 (3.3)
87.7 (2.7)
87.6 (2.9)
76.9 (6.6)
92.5 (1.8)
84.8 (5.9)
80.1 (4.1)
88.9 (2.4)
93.0 (2.0)
92.4 (0.6)
91.1 (3.0)
71.9 (2.4)
90.4 (2.0)
90.1 (1.1)

107 (2.4)
106 (2.8)
105 (2.4)
The number of determinations was 3.
the average recoveries.
The values given in parentheses are the percent relative standard deviations of
                                                     8061 - 17
                                                                      Revision 0
                                                                  September 1994

-------
                                                          TABLE 5.
                                ACCURACY AND PRECISION DATA FOR METHOD 3550 AND METHOD 8061a
Spike Concentration
(1 mq/kq)
Compound
Dimethyl phthalate
Diethyl phthalate
Diisobutyl phthalate
Di-n-butyl phthalate
Bis(4-methyl-2-pentyl) phthalate
Bis(2-methoxyethyl) phthalate
Diamyl phthalate
Bis(2-ethoxyethyl) phthalate
Hexyl 2-ethylhexyl phthalate
Dihexyl phthalate
Benzyl butyl phthalate
Bis(2-n-butoxyethyl ) phthalate
Bis(2-ethylhexyl) phthalate
Dicyclohexyl phthalate
Di-n-octyl phthalate
Dinonyl phthalate
Estuarine
sediment
77.9
68.4
103
121
108
26.6
95.0
C
c
103
113
114
C
36.6
C
c
(42.8)
(1.7)
(3.1)
(25.8)
(57.4)
(26.8)
(10.2)


(3.6)
(12.8)
(21.1)

(48.8)


Municipal
sludge
52.1
68.6
106
86.3
97.3
72.7
81.9
66.6
114
96.4
82.8
74.0
76.6
65.8
93.3
80.0
(35.5)
(9.1)
(5.3)
(17.7)
(7.4)
(8.3)
(7.1)
(4.9)
(10.5)
(10.7)
(7.8)
(15.6)
(10.6)
(15.7)
(14.6)
(41.1)
Sandy loam
soil
c
54.7
70.3
72.6
c
0
81.9
c
57.7
77.9
56.5
c
99.2
92.8
84.7
64.2

(6.2)
(3-7)
(3.7)


(15.9)

(2.8)
(2.4)
(5-1)

(25.3)
(35.9)
(9.3)
(17.2)
Spike Concentration
(3 uq/q)
Estuarine
sediment
136
60.2
74.8
74.6
104
19.5
77.3
21.7
72.7
75.5
72.9
38.3
59.5
33.9
36.8
c
(9.6)
(12.5)
(6.0)
(3.9)
(1.5)
(14.8)
(4.0)
(22.8)
(11.3)
(6.8)
(3.4)
(25.1)
(18.3)
(66.1)
(16.4)

Municipal
sludge
64.8
72.8
84.0
113
150
59.9
116
57.5
26.6
80.3
76.8
98.0
85.8
68.5
88.4
156
(11.5)
(10.0)
(4.6)
(5.8)
(6.1)
(5.4)
(3-7)
(9.2)
(47.6)
(4.7)
(10.3)
(6.4)
(6.4)
(9.6)
(7.4)
(8.6)
Sandy loam
soil
70.2 (2.0)
67.0 (15.1)
79.2 (0.1)
70.9 (5.5)
83.9 (11.8)
0
82.1 (15.5)
84.7 (8.5)
28.4 (4.3)
79.5 (2.7)
67.3 (3.8)
62.0 (3.4)
65.4 (2.8)
62.2 (19.1)
115 (29.2)
115 (13.2)
a  The number of determinations was 3.  The values given in parentheses are the percent relative standard deviations of the
   average recoveries.  All samples were subjected to Florisil cartridge cleanup.

b  The estuarine sediment extract (Florisil, Fraction 1) was subjected to sulfur cleanup (Method 3660 with
   tetrabutylammonium sulfite reagent).

c  Not able to determine because of matrix interferant.
                                                          8061 - 18
    Revision 0
September 1994

-------
                                     Figure 1
                                                                OB-5
                                                                30 m x 0.53 mm 10
                                                                1.S~um Rim
                          IS
             11   12 SU-1 SU-2  SU-3
                     v«w_*A'
                                                                     16
     O
     LU
                u
                       IS
JL
                                        10
                                DB-1701
                     SU-2 SU-3   30 m x 0.53 mm ID
                        II      1 0— urn Film
             12  SU-1 15  1 T
              13
                     10
    20

TIME (min)
                                                      30
                                      40
GC/ECD chromatograms of a composite  phthalate esters standard  (concentration
10 ng//iL  per compound) analyzed  on  a DB-5 and a DB-1701  fused-silica open
tubular column.   Temperature program:  150°C  (0.5 min hold) to 220°C at
5°C/min,  then to  275°C  (13 min  hold)  at 3°C/nnn.
                                    8061 -  19
                                      Revision 0
                                  September 1994

-------
                                 METHOD  8061
    PHTHALATE  ESTERS  BY  CAPILLARY  GAS  CHROMATOGRAPHY
          WITH  ELECTRON  CAPTURE DETECTION  (GC/ECD)
        7 1 Extraction
711 Refer to Chapter 2 for
     guidance on choosing
     an extraction procedure.
     Recommendations given.
7.1.2 Determine spike sample
     recovery and detection limit
     for each new sample matrix
     and a given extraction
     procedure.
713 Aqueous sample extraction
     with C18 disks:
      1 Precondition disks using
       solvent train.
      2 Concentrate sample
       analytes on disk.
      3 Elute sample analytes
       with acetonitrile.
     .4 Concentrate extract:
        1 Micro-Snyder Column
         Technique
        2 Nitrogen Slowdown
         Technique
          1 Evaporate solvent to
           desired level
         2 Rinse tube walls
           frequently and avoid
           evaporating todryness.
             I
 7 2 Solvent Exchange to Hexane
 7 2.1 Evaporate extract volume to
 1 -2 ml using K-D assembly
7 2.2 Add hexane to K-D assembly
and evaporate to 1-2 mL
 7 23 Rinse K-D components and
 adjust volume to desired level.
                                                                        7 3 Cleanup/FracoonaOon
 73.1 Cleanup may not be
      necessary tor extracts with
      clean sample matrices
      Fraction collection and
      methods outlined tor other
      compd groups of interest.
7.3.2 Flonsil Cartridge Cleanup
     1 Check each lot of Rorisil
       cartridges for analyte
       recovery by eluting and
       analyzing a composite std
     2 Wash and adjust solvent
       flow through cartridges.
     3 Place culture tubes or 5 mL
       vol. flasks for eluate
       collection.
     4 Transfer appropriate extract
       volume on cartridge
     5 Elute the cartridges and
       dilute to mark on flask
       Transfer eluate to glass
       vials for concentration.
733 Collect 2 fractions it PCBs
     and organochlorme pesticides
     are known to be present
     7 4 Gas Chromatograph
                                                                    7 41 Set GC operating parameters
                                                                    7.4 2 Table 1 and Figure 1 show
                                                                         MOLs and analyte retention
                                                                         times.
                                   8061  -  20
                    Revision  0
              September  1994

-------
     METHOD  8061
     (CONTINUED)
         9
        75 Calibration
 751  See Method 8000 for
      calibration technique.
 7.52 Refer to Method 8000 tor
      internal/external std.
      procedure.
       7 6 GC Analysis
  7 6 i Refer (o Memod 8000
 762 Follow Section 76 in
     Method 8000 for
     instructions on analysis
     sequence, dilutions.
     retention time windows.
     and identification criteria
 7 6.3  Record injection volume
      and sample peak areas
7 64 Identify and quantify each
     component peak using the
     internal or external std
     procedure.
765 Dilute extracts which
     show analyte levels
     outside of the calibration
     range.
766 Identify compounds in the
     sample by comparing
     retention times in the
     sample and the standard
     chromatograms
      8061  -  21
      Revi sion  0
September  1994

-------
00
o

-------
                                 METHOD 8070

                      NITROSAMINES BY GAS CHROMATQGRAPHY
1.0   SCOPE AND APPLICATION

      1.1   This method covers the determination of certain nitrosamines.   The
following compounds can be determined by this method:


                                                 Appropriate Technique
Compound Name               CAS No.a    3510    3520    3540    3550    3580
N-Nitrosodimethylamine 62-75-9 X X X X
N-Nitrosodiphenylamine 86-30-6 X X X X
N-Nitrosodi-n-propylamine 621-64-7 X X X X
X
X
X
a Chemical Abstract Services Registry Number.
X Greater than 70 percent recovery by this preparation technique.
      1.2   This  is  a  gas  chromatographic  (GC)  method  applicable  to  the
determination  of the  parameters  listed  above  in  municipal  and  industrial
discharges.  When this method is used to analyze unfamiliar samples for any or
all of the compounds above, compound identifications  should be supported by at
least one  additional qualitative technique.   This  method  describes analytical
conditions for a second gas chromatographic column that can be used to confirm
measurements  made  with  the  primary  column.    Method   8270  provides  gas
chromatograph/mass  spectrometer   (GC/MS)   conditions  appropriate   for  the
qualitative    and    quantitative    confirmation    of    results    for
N-nitrosodi-n-propylamine.      In    order   to  confirm   the   presence   of
N-nitrosodiphenylamine, the cleanup procedure specified in Section 7.3.3 or 7.3.4
must be  used.   In order to confirm  the presence  of  N-nitrosodimethylamine by
GC/MS, chromatographic column 1 of this method must be substituted for the column
recommended  in  Method  82-70.   Confirmation  of these  parameters  using GC-high
resolution mass  spectrometry or a  Thermal  Energy  Analyzer is  also recommended
practice.

      1.3   The  method  detection  limit (MDL)  for each parameter is  listed in
Table 1.   The MOL  for a  specific  wastewater may differ from  those listed,
depending upon the nature of  interferences  in  the sample matrix.  Table 2 lists
the Estimated Quantitation Limits (EQLs) for various matrices.

      1.4   The toxicity or carcinogenicity of each reagent used  in this method
has not  been precisely  defined.   However,  each  chemical compound  should be
treated  as a  potential  health hazard.   From this  viewpoint,  exposure to these
chemicals must be reduced  to the lowest possible concentration  by  whatever means
available.  The  laboratory is responsible  for maintaining a current awareness
file of OSHA regulations regarding  the safe  handling  of the chemicals specified
in this  method.  A reference file of material data handling sheets should also

                                   8070 - 1                         Revision 0
                                                                     July 1992

-------
be made available to all personnel  involved in the chemical  analysis.

      1.5   These nitrosamines are known carcinogens.   Therefore,  utmost care
must be exercised  in  the handling  of these materials.   Nitrosamine reference
standards and standard solutions  should  be  handled and prepared in a ventilated
glove box within a properly ventilated room.

      1.6   N-Nitrosodiphenylamine   is  reported  to  undergo  transnitrosation
reactions.  Care must  be exercised  in  the heating or concentrating of solutions
containing this compound in the presence of reactive amines.


2.0   SUMMARY OF METHOD

      2.1   A measured  volume  of aqueous  sample, approximately  one liter,  is
solvent  extracted  with  methylene  chloride using  a  separatory  funnel.   The
methylene chloride  extract  is  washed with dilute  HC1 to remove free  amines,
dried, and  concentrated to a  volume  of 10 mi  or less.  Gas  chromatographic
conditions  are  described which permit  the separation and  measurement  of the
compounds in the extract after it has been exchanged to methanol.

      2.2   Method 8070 provides gas chromatographic conditions for the detection
of ppb concentrations  of nitrosamines.  Prior to use of this method, appropriate
sample extraction techniques must be used. •  Both neat and diluted.organic liquids
(Method 3580, Waste Dilution) may be  analyzed  by direct  injection.  A 2 to 5 pi
aliquot of  the  extract  is  injected  into  a gas  chromatograph (GC) using the
solvent flush technique,  and  compounds in  the  GC effluent  are  detected by a
nitrogen-phosphorus detector (NPD) or a Thermal Energy Analyzer and the reductive
Hall detector.


3.0   INTERFERENCES

      3.1   Refer to Methods 3500,  3600, and 8000.

      3.2   Matrix  interferences  may  be   caused  by   contaminants  that  are
coextracted  from the   sample.   The extent  of matrix  interferences will vary
considerably from source to source, depending upon the nature and diversity of
the industrial  complex  or municipality  being  sampled.   The  cleanup procedures
(Methods 3610 or 3620) can be used  to  overcome many of these interferences, but
unique samples  may require additional  cleanup  approaches  to  achieve  the MDL
listed in Table  1.

      3.3   Nitrosamines contaminate many types of products commonly found in the
laboratory.    The   analyst  must  demonstrate that  no  nitrosamine  residues
contaminate  the sample or solvent extract under the  conditions of analysis.
Plastics, in particular, must be  avoided because nitrosamines are commonly used
as  plasticizers and  are easily extracted from  plastic materials.   Serious
nitrosamine contamination may result at any time if consistent quality control
is not practiced.

      3.4   The sensitive and selective Thermal Energy Analyzer and the reductive
Hall detector may  be   used  in  place  of the nitrogen-phosphorus  detector when

                                   8070 -  2                         Revision 0
                                                                     July 1992

-------
interferences are encountered.  The Thermal Energy Analyzer offers tha highest
selectivity of the non-mass spectrometric detectors.

      3.5   Solvents, reagents,  glassware,  and other sample processing hardware
may yield discrete artifacts and/or elevated baselines causing misinterpretation
of gas chromatograms.  All  these materials  must be demonstrated to be free from
interferences, under the conditions of the analysis, by analyzing reagent blanks.
Specific selection of reagents and purification of solvents by distillation in
all-glass systems may be required.

      3.6   Interferences coextracted from samples will vary considerably from
source to  source,  depending upon  the  waste being  sampled.   Although  general
cleanup techniques are recommended as  part of  this  method,  unique samples may
require additional  cleanup.


4.0   APPARATUS AND MATERIALS

      4.1   Gas chromatograph - An analytical system complete with temperature
programmable gas chromatograph suitable for on-column injection and all required
accessories including syringes,  analytical  columns,  gases, detector, and strip-
chart recorder.  A data system is recommended for measuring peak areas.

            4.1.1 Column !-1.8mx4mmID Pyrex glass, packed with Chromosorb
      W AW, (80/100  mesh)  coated  with  10%  Carbowax  20  M/2% KOH  or equivalent.
      This  column  was used  to  develop the  method  performance  statements  in
      Section 9.0.   Guidelines for  the  use of alternate column  packings are
      provided in Section 7.3.2.

            4.1.2 Column  2  -  1.8 m  x 4  mm  ID  Pyrex  glass,  packed  with
      Supelcoport (100/120 mesh) coated with 10% SP-2250, or equivalent.

            4.1.3 Detector  -  Nitrogen-Phosphorus,  reductive  Hall  or  Thermal
      Energy Analyzer. These  detectors have proven effective in the analysis of
      wastewaters for the parameters  listed in the scope.  A nitrogen-phosphorus
      detector was used to  develop the method performance statements in Section
      9.0.   Guidelines  for  the  use  of alternate  detectors  are  provided  in
      Section 7.3.2.

      4.2   Kuderna-Danish (K-D) apparatus

            4.2.1 Concentrator tube - 10 mL,  graduated  (Kontes K-570050-1025 or
      equivalent).  Calibration must be checked at the volumes employed in the
      test.  A ground glass stopper is used to prevent evaporation of extracts.

            4.2.2 Evaporation   flask  -  500  mL   (Kontes   K-570001-0500  or
      equivalent).    Attach  to  concentrator tube  with  springs,   clamps,  or
      equivalent.

            4.2.3 Snyder column - Three ball  macro (Kontes  K-503000-0121  or
      equivalent).

            4.2.4 Snyder  column  -  Two  ball micro (Kontes  K-569001-0219  or

                                   8070 -  3                         Revision 0
                                                                     July 1992

-------
      equivalent).

            4.2.5 Springs -  1/2 inch (Kontes K-662750 or equivalent).

      4.3   Boiling chips - Approximately 10/40 mesh.  Heat to 400°C for
30 minutes or Soxhlet extract with methylene chloride.

      4.4   Water  bath  -  Heated,  with  concentric  ring  cover,  capable  of
temperature control (± 2°C).   The bath should be used in a  hood.
top.
      4.5   Balance - Analytical, 0.0001 g.

      4.6   Vials - 10 to 15 ml,  amber glass with Teflon lined screw-cap or crimp
      4.7   Volumetric  flasks,  Class A, Appropriate  sizes with  ground  glass
stoppers.


5.0   REAGENTS

      5.1   Reagent grade inorganic chemicals shall  be used in all  tests.  Unless
otherwise indicated, it is intended that all inorganic reagents shall conform to
the  specifications  of  the  Committee on Analytical  Reagents of  the  American
Chemical Society, where such specifications are  available.  Other grades may be
used, provided it is first ascertained that the  reagent is of sufficiently high
purity to permit its use without lessening the accuracy of the determination.

      5.2   Organic-free reagent water - All references to water  in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.3   Methanol, CH3OH - Pesticide quality or equivalent.

      5.4   Isooctane,  (CH3)3CCH2CH(CH3}2 -  Pesticide quality or  equivalent.

      5.5   Methylene chloride, CH2C12 - Pesticide  quality  or equivalent.

      5.6   Stock standard solutions (1000  mg/L) -  Stock standard solutions can
be prepared from pure standard materials or purchased as certified solutions.

            5.6.1 Prepare  stock  standard   solutions  by  accurately  weighing
      0.1000 ± 0.0010 g of pure  material.   Dissolve the  material in pesticide
      quality  methanol  and  dilute  to  volume  in  a  100  ml  volumetric  flask.
      Larger volumes can be used  at the convenience of the analyst.  If compound
      purity  is  certified at 96% or  greater,  the weight can be used without
      correction  to  calculate  the  concentration  of  the  stock  standard.
      Commercially prepared stock standards can be  used at any concentration if
      they are certified  by the manufacturer or by an independent source.

            5.6,2 Transfer the stock standard solutions into bottles with Teflon
      lined screw-caps  or crimp tops.   Store at 4°C and protect  from light.
                                   8070 - 4                         Revision 0
                                                                     July 1992

-------
      Stock  standard solutions  should  be  checked  frequently  for  signs  of
      degradation or evaporation, especially just prior to preparing calibration
      standards from them.

            5.6.3 Stock standard solutions must be replaced after six months, or
      sooner if comparison with check standards indicates a problem.

      5.7   Calibration standards - A minimum of five concentrations  should be
prepared through dilution  of the stock standards with  isooctane.  One  of the
concentrations should be at a concentration near, but above, the method detection
limit.  The remaining concentrations  should correspond  to the expected range of
concentrations found in real samples or should define the working range of the
GC.   Calibration  solutions  must  be  replaced after  six months, or  sooner if
comparison with check standards indicates a problem.

      5.8   Internal standards (if internal  standard calibration is used) - To
use this approach, the analyst must select one or more internal standards that
are similar in analytical  behavior to  the compounds  of interest.  The analyst
must further demonstrate that  the measurement of  the internal  standard  is not
affected by method  or matrix interferences.   Because of these limitations, no
internal standard can be suggested that is applicable to all samples.

            5.8.1 Prepare   calibration  standards   at   a  minimum   of  five
      concentrations for each analyte of interest, as described  in Section 5.7.

            5.8.2 To each calibration standard,  add a known constant  amount of
      one or more internal standards, and dilute to volume with isooctane.

            5.8.3 Analyze each calibration standard according to Section 7.0.

      5.9   Surrogate standards - The analyst  should  monitor the performance of
the extraction, cleanup  (when used), and analytical  system and the effectiveness
of  the  method  in  dealing  with  each  sample  matrix  by spiking each  sample,
standard, and reagent blank with one or two surrogates (e.g. nitrosamines that
are not expected to be in  the sample) recommended to  encompass the range of the
temperature program used in this method. Method 3500 details instructions on the
preparation of base/neutral surrogates.  Deuterated analogs of analytes should
not be  used as surrogates  for gas  chromatographic  analysis due  to  coelution
problems.


6.0   SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1   See the introductory material  to this chapter,  Organic  Analytes,
Section  4.1.   Extracts  must be  stored at  4°C and analyzed within 40  days of
extraction.
                                   8070 - 5                         Revision 0
                                                                     July 1992

-------
7.0   PROCEDURE

      7.1   Extraction

            7.1.1 Refer to Chapter Two  for guidance on choosing the appropriate
      extraction procedure.    In  general,  water  samples are  extracted at  a
      neutral, or as  is, pH with methylene chloride, using either Method 3510 or
      3520.  Solid  samples are extracted using either Method 3540 or 3550.

            7.1.2 Prior to gas chromatographic analysis, the extraction solvent
      must be exchanged to methanol.   The exchange is performed during the  K-D
      procedures listed  in all of the extraction methods.   The  exchange  is
      performed as  follows.

                  7.1.2.1     Following K-D of the methylene  chloride extract to
      1 ml using the  macro-Snyder  column, allow the apparatus to cool and drain
      for at least  10 minutes.

                  7.1.2.2     Momentarily remove the Snyder column, add 50 ml of
            methanol, a new boiling chip, and reattach the macro-Snyder column.
            Concentrate the extract using 1 ml of methanol to prewet the Snyder
            column.   Place the K-D  apparatus  on the water bath so  that  the
            concentrator tube  is  partially immersed  in the  hot  water.  Adjust
            the vertical  position  of the apparatus and the water temperature, as
            required, to complete  concentration in 5-10 minutes.  At the proper
            rate of distillation the  balls of the column will actively chatter,
            but the chambers will  not flood.  When the apparent volume of liquid
            reaches  1 ml,  remove  the K-D apparatus and allow  it  to drain  and
            cool for at  least  10  minutes.    The  extract will  be  handled
            differently at  this point,  depending on  whether or not cleanup is
            needed.  If cleanup is  not required,  proceed to Section 7.1.2.3.   If
            cleanup  is needed, proceed to Section 7.1.2.4.

                  7.1.2.3      If cleanup of the extract is not required, remove
            the Snyder column  and rinse the flask and its lower joint into  the
            concentrator  tube with  1-2  ml  of methanol.   A  5  ml  syringe  is
            recommended  for  this  operation.   Adjust the  extract volume  to
            10.0 ml.   Stopper the  concentrator tube  and  store  refrigerated at
            4°C if  further processing will not be  performed immediately.  If the
            extract  will  be  stored  longer  than   two  days,  it  should   be
            transferred to  a  vial  with  a Teflon  lined screw-cap or crimp top.
            Proceed with gas chromatographic  analysis.

                  7.1.2.4      If cleanup of the extract is required, remove  the
            Snyder column and rinse the  flask and  its  lower joint  into  the
            concentrator tube with a  minimum  amount of methylene chloride.  A 5
            ml syringe is recommended for this operation.  Add a clean boiling
            chip to  the  concentrator tube  and attach a  two ball  micro-Snyder
            column.   Prewet  the  column by  adding about  0.5 ml  of methylene
            chloride to the top.   Place the  micro K-D apparatus  on the water
            bath (80°C)  so that the concentrator tube is  partially immersed in
            the hot  water.  Adjust the  vertical  position  of the apparatus  and
            the water temperature, as required, to complete concentration in 5-

                                   8070 - 6                         Revision 0
                                                                     Ouly 1992

-------
      10 minutes.   At the proper rate of  distillation  the balls of the
      column will actively chatter,  but the chambers will not flood.  When
      the  apparent  volume  of  liquid  reaches  0.5 ml,  remove  the  K--D
      apparatus and  allow it to drain and cool for at least 10 minutes.

            7.1.2.5     Remove the micro-Snyder column and rinse the flask
      and  its  lower  joint  into  the  concentrator tube  with 0.2 ml  of
      methylene chloride.   Adjust  the extract volume to 2.0 ml and proceed
      with either Method 3610, 3620, or 3640.

      7.1.3 If   N-nitrosodiphenylamine   is   to   be  measured   by   gas
chromatography, the  analyst must  first use a cleanup column to eliminate
diphenylamine   interference   (Methods    3610   or   3620).      If   N-
nitrosodiphenylamine  is of no interest,  the analyst may proceed directly
with gas chromatographic analysis (Section 7.3).

7.2   Cleanup

      7.2.1 Cleanup procedures may not be necessary for a relatively clean
sample matrix.   The cleanup procedure recommended in this method has been
used for the analysis of various clean waters and  industrial effluents. If
particular  circumstances  demand   the  use  of  an  alternative  cleanup
procedure, the analyst must determine the elution profile and demonstrate
that  the recovery  of each compound of  interest  is  no less  than  85%.
Diphenylamine, if present in the original  sample  extract must be separate
from  the nitrosatnines if N-nitrosodiphenylamine is to  be determined by
this method.

      7.2.2 Proceed  with  either  Method  3610  or 3620,  using  the  2  ml
methylene chloride extracts obtained from Section 7.1.2.5.

      7.2.3 Following cleanup, the extracts should be analyzed by GC, as
described in the previous paragraphs and in Method 8000.

7.3   Gas Chromatography

      7.3.1 N-nitrosodiphenylamine completely  reacts to form diphenylamine
at the normal operating temperatures of a GC injection port (200 to 250°C).
Thus,   N-nitrosodiphenylamine  is  chromatographed   and   detected   as
diphenylamine.  Accurate determination depends on removal of diphenylamine
that  may be present  in  the original  extract prior to GC (see  Section
7.1.3).

      7.3.2 Table  1  summarizes  the  recommended operating conditions for
the gas chromatograph.  This table includes retention  times and MOLs that
were  obtained  under  these  conditions.    Examples  of  the  parameter
separations achieved by these columns are shown in Figures 1 and 2.  Other
packed columns, chromatographic conditions, or detectors  may be  used if
the requirements of Section 8.2 are met.  Capillary  (open-tubular) columns
may  also be used  if  the  relative standard deviations  of responses for
replicate  injections  are  demonstrated  to  be  less   than  6%  and  the
requirements of Section 8.2 are met.


                             8070 - 7                         Revision 0
                                                               July 1992

-------
      7.4   Calibration - Refer to Method 8000 for proper calibration techniques.
Use Table 1  and especially Table 2 for guidance on selecting the lowest point on
the calibration curve.

            7.4.1 The  procedure  for internal  or external  calibration  may  be
      used.   Refer to  Method 8000 for a description of each of these procedures.

            7.4.2 If  cleanup  is  performed  on the  samples, the  analyst  should
      process a  series of  standards  through  the  cleanup  procedure  and  then
      analyze the  samples by  GC.   This will  confirm elution  patterns and the
      absence of interferents from the reagents.

      7.5   Gas chromatographic analysis

            7.5.1 Refer to Method 8000.   If the internal  standard calibration
      technique is used, add 10 pi of internal standard to the sample prior to
      injection.

            7.5.2 Method 8000 provides instructions on  the  analysis sequence,
      appropriate  dilutions,  establishing  daily  retention  time  windows,  and
      identification criteria.  Include a mid-concentration check standard after
      each group of 10 samples in the analysis sequence.

            7.5.3 Examples of GC/NPD chromatograms for nitrosamines are shown in
      Figures 1 and 2.

            7.5.4 Record the sample  volume injected and  the resulting peak sizes
      (in area units or peak heights).

            7.5.5 Using either the  internal  or external  calibration procedure
      (Method 8000), determine the identity and quantity of each analyte peak in
      the sample chromatogram.  See Method 8000 for calculation equations.

            7.5.6  If  peak  detection and  identification  are prevented  due  to
      interferences, the hexane extract may  undergo cleanup using either Method
      3610 or 3620.
8.0   QUALITY CONTROL

      8.1   Refer  to Chapter  One  for specific  quality control  procedures.
Quality control to validate sample extraction is covered in Method 3500 and in
the extraction method utilized.   If extract cleanup was  performed,  follow the QC
in Method 3600 and in the specific cleanup method.

      8.2   Procedures to check the GC system operation are found in Method 8000,
Section 8.6.

            8.2.1 The quality control  (QC)  reference  sample concentrate (Method
      8000, Section 8.6) should  contain each analyte of interest at 20 mg/L.

            8.2.2 Table 3 indicates the calibration and QC acceptance criteria
      for this method.  Table 4 gives method accuracy and precision as functions

                                   8070 - 8                         Revision 0
                                                                     July 1992

-------
      of concentration  for the analytes  of interest.   The contents of  both
      Tables should be used to evaluate a  laboratory's  ability  to  perform and
      generate acceptable data by this method.

      8.3   Calculate surrogate standard recovery  on  all  samples,  blanks,  and
spikes.  Determine if the  recovery is within  limits  (limits  established  by
performing  QC procedures outlined in Method 8000, Section 8.10).

            8.3.1 If recovery is not within limits, the following is required.

            •     Check to  be  sure  that  there are no errors  in calculations,
                  surrogate  solutions  and  internal   standards.   Also,  check
                  instrument performance.

            •     Recalculate the data and/or reanalyze the extract if  any of
                  the above checks reveal  a problem.

            •     Reextract and reanalyze  the  sample if  none of  the above are a
                  problem or flag the data as  "estimated concentration.


9.0   METHOD PERFORMANCE

      9.1   This method has  been  tested for linearity of recovery  from  spiked
organic-free reagent water  and has  been demonstrated  to be applicable for the
concentration range from 4 x MDL to 1000 x MOL.

      9.2   In a single laboratory (Southwest Research Institute), using spiked
wastewater samples, the average recoveries presented in Table 2 were obtained.
Each spiked sample was analyzed in triplicate on  three separate  occasions.  The
standard deviation of the percent recovery is also included in  Table 2.


10.0  REFERENCES

1.    Fed.  Regist. 1984, 49, 43234; October 26.

2.    "Determination of Nitrosamines in Industrial and Municipal Wastewaters";
      Report for EPA Contract 68-03-2606,  in preparation.

3.    Burgess,   E.M.;  Lavanish,   J.M.   "Photochemical   Decomposition  of  N-
      nitrosamines"; Tetrahedron  Letters 1964, 1221.

4.    Methods  for  Chemical Analysis  of Water and Wastes:  U.S. Environmental
      Protection  Agency.  Office  of  Research  and Development. Environmental
      Monitoring and Support Laboratory. ORD Publication Offices of Center for
      Environmental Research Information:  Cincinnati,  OH,  1979; EPA-600/4-79-
      020.

5.    "Method Detection Limit and  Analytical Curve Studies EPA Methods  606, 607,
      608"; U.S. Environmental Protection Agency.  Environmental Monitoring and
      Support Laboratory, Cincinnati,  OH,  special letter report for EPA Contract
      68-03-2606.

                                    8070 - 9                         Revision 0
                                                                     July 1992

-------
                                   TABLE 1.
            CHROMATOGRAPHIC CONDITIONS AND METHOD DETECTION LIMITS
Analyte
  Retention Time
     (minutes)	
Column 1    Column 2
                                                                Method
                                                            Detection Limit
N-Ni trosodimethyl ami ne
N-Nitrosodi-n-propyl amine
N-Ni trosodi phenyl ami ne*
4.1
12.1
12. 8b
0.88
4.2
6.4C
0.15
0.46
0.81
Column 1 conditions:
   Carrier gas (He) flow rate:
   Column temperature:
Column 2 conditions:
   Carrier gas (He) flow rate:
   Column temperature:
  40 mL/min
  Isothermal
  indicated.
  40 mL/min
  Isotherma.1
  indicated.
at  110°C,   except  as   otherwise
at  120°C,   except  as   otherwise
a  Measured as diphenylamine.
b  Determined isothermally at 220°C.
c  Determined isothermally at 210°C.
                                   TABLE 2.
                    SINGLE OPERATOR ACCURACY AND PRECISION
                              Average   Standard    Spike
                              Percent   Deviation   Range
                              Number
                                of     Matrix
Analyte
Types
N-Nitrosodimethyl amine
N-Nitrosodiphenyl amine
N-Ni trosodi -n-propyl ami ne
Recovery % (MgA)

32
79
61

3.7
7.1
4.1

0.8
1.2
9.0
Analyses

29
29
29

5
5
5
                                   8070 -  10

-------
                                   TABLE 3.
                            QC ACCEPTANCE CRITERIA
                                   Test      Limit       Range      Recovery
                                   Cone.      for s       for X       Range
 Analyte                           (jig/L)
 N-Nitrosodimethylamine              20        3.4       4.6-20.0     13-109

 N-Nitrosodiphenylamine              20        6.1       2.1-24.5      D-139

 N-Nitrosodi-n-propylamine           20        5.7      11.5-26.8     45-146


s     =     Standard deviation for four recovery measurements,  in pg/L.

X     =     Average recovery for four recovery measurements,  in

D     =     Detected, result must be greater than zero.
                                   8070 - 11                        Revision 0
                                                                     July  1992

-------
                                   TABLE 4.
         METHOD  ACCURACY  AND  PRECISION  AS  FUNCTIONS OF  CONCENTRATION
Analyte
N-Ni trosodimethyl ami ne
N-Ni trosodi phenyl ami ne
N-Ni troso-n-propyl ami ne
Accuracy, as
recovery, X'
(W/U
0.37C+0.06
0.64C+0.52
0.96C-0.07
Single
analyst
precision,
sr' (ng/L)
0.25X-0.04
0.36X-1.53
0.15X+0.13
Overall
precision,
S' (ng/L)
0.25X+0.11
0.46X-0.47
0.21X+0.15
c

X
Expected  recovery  for  one  or  more  measurements  of  a  sample
containing a concentration of C,  in [ig/L.

Expected single  analyst  standard  deviation of measurements  at  an
average concentration found of X,  in jig/L.

True value for the concentration,  in ng/L.

Average recovery found for measurements  of samples containing  a
concentration of C, in yg/L.
                                   8070 - 12
                                                        Revision 0
                                                         July 1992

-------
             FIGURE  1.
GAS CHROMATOGRAM OF NITROSAMINES
    Column: 10% Csrbowtt 20M + 2%
           KOH on Chromosorb W-AW
    Timptftturt: 7/0°
    Dittctor: Phosphorus/Nitrogin
          _J	I    I   I
      2   4   S  8   10  12  !4

                     minute*
            8070  -  13
Revision 0
 July  1992

-------
                           FIGURE 2.
GAS CHROMATOGRAM OF N-NITROSODIPHENYLAMINE AS  DIPHENYLAMINE
               Column- IO% CtrbowtJt 20M * 2% KQH on
                      Chromosorb W-4W
               remptnturi: 220° C.
               Dtttctor Phosphorus/Nitrogen
                                          Q
              0   2  4   6   8   tO   12  14  16 19

                                 tim*. minvt**
                            8070 - 14
Revision  0
 July  1992

-------
                                 METHOD  8070
                 NITROSAMINES BY  GAS CHROMATOGRAPHY
                           7 1  i Chaota
                            apprpri«la
                            an t. raclion
                             procadura
                           7  1  2 P«cfoc»
                          olvanl ««cnan9a
                          u»in^ malhanol
  '124 P.rfor.
•icro-K-D procadura
  uiing mathylan*
 chlorida  p«r(ar«
  Malhod 3610 or
3620.  pcoca»d »iIh
    CC analyai.1

7
I 2 3
Ho
Adjuit
procaad »
anal yaia or
nannar
th
• to r*

   7  1  : P.rfor.
  column cleanup
 utino,  Method 3610
      or 3620
  1  3  2 Rafar  U
    Tibia 1 for
    raeo*aandad
    oparatin?
condition* for tha
       CC
  4 RaCar to  Malhod
  3000 for propar
   calibra Lion
   Lachniqua*
                                                  7 $ 1 Rafar to
                                                  .lalhod 9000 for
                                                  9uidanca  on CC
                                                    inalyt11
                                                 5 V 5  S Raeord
                                                  tav>pla  vo 1 u*a
                                                  in]aelad and
                                                 fa«ulting paak
                                                  ilia/parforn
                                                   •ppcopriala
                                                alcuUlioni Irafar
                                                 to Malhod 800G)
                                                      Stop
                                    8070  -  15
                                   Revision  0
                                    July  1992

-------
00
©
00
o

-------
                                 METHOD 8080

                     ORGANOCHLORINE PESTICIDES AND PCBs
1.0  SCOPE AND APPLICATION

     1.1  Method 8080  1s  used  to  determine  the  concentration  of various
organochlorine pesticides and  polychlorinated  biphenyls  (PCBs).     Table 1
indicates compounds that may be determined by this method and lists the method
detection limit for  each  compound  in  reagent  water.    Table  2 lists the
practical quantitation limit (PQL) for other matrices.


2.0  SUMMARY OF METHOD

     2.1  Method  8080  provides   gas   chromatograpMc  conditions  for  the
detection of ppb levels of certain  organochlorine pesticides and PCBs.  Prior
to the use of this  method,  appropriate  sample extraction techniques must be
used.  Both neat and diluted organic liquids (Method 3580, Waste Dilution) may
be analyzed by direct Injection.  A  2-  to 5-uL sample is injected into a gas
chromatograph  (GC) using the solvent flush  technique, and compounds in the GC
effluent are detected by  an  electron  capture  detector  (ECD) or a halogen-
specific detector  (HSD).

     2.2  The  sensitivity of  Method  8080  usually  depends  on  the level of
Interferences  rather  than  on  Instrumental  limitations.    If  Interferences
prevent  detection  of  the  analytes,  Method  8080  may  also  be performed on
samples  that have  undergone cleanup.  Method 3620, Flor1s1l Column Cleanup, by
itself or followed by Method  3660,  Sulfur  Cleanup, may be used to eliminate
interferences  in the analysis.


3.0   INTERFERENCES

      3.1 Refer to Methods 3500  (Section  3.5, in particular), 3600, and 8000.

      3.2 Interferences  by  phthalate   esters   can  pose  a  major  problem 1n
pesticide determinations when   using   the electron   capture  detector.   These
compounds generally  appear 1n   the  chromatogram   as  large late-eluting peaks,
especially  in  the  15% and 50%   fractions from the  Florisll cleanup.  Common
flexible plastics  contain varying amounts of phthalates.  These  phthalates  are
easily  extracted or  leached  from  such  materials during  laboratory operations.
Cross contamination  of   clean   glassware   routinely   occurs  when plastics  are
handled during extraction steps,   especially  when solvent-wetted  surfaces  are
handled.  Interferences   from   phthalates can  best  be minimized by  avoiding
contact with   any  plastic  materials.     Exhaustive  cleanup  of reagents  and
glassware may  be   required   to   eliminate background phthalate  contamination.
The  contamination  from  phthalate  esters   can   be  completely eliminated with a
microcoulometrlc or  electrolytic conductivity detector.
                                   8080 - 1
                                                          Revision       0
                                                          Date   September  1986

-------
TABLE 1.  GAS CHROMATOGRAPHY OF PESTICIDES AND PCBsa
Retention time (m1n)

Compound
Aldrln
a-BHC
£-BHC
ff-BHC
7-BHC (Llndane)
Chlordane (technical)
4,4'-DDD
4,4 '-DDE
4,4'-DDT
D1eldr1n
Endosulfan I
Endosulfan II
Endosulfan sulfate
Endrln
Endrln aldehyde
Heptachlor
Heptachlor epoxlde
Methoxychlor
Toxaphene
PCB-1016
PCB-1221
PCB-1232
PCB-1242
PCB-1248
PCB-1254
PCB-1260
aU.S. EPA. Method
Environmental Monitoring and

Col. 1
2.40
1.35
1.90
2.15
1.70
e
7.83
5.13
9.40
5.45
4.50
8.00
14.22
6.55
11.82
2.00
3.50
18.20
e
e
e
e
e
e
e
e
617.

Col. 2
4.10
1.82
1.97
2.20
2.13
e
9.08
7.15
11.75
7.23
6.20
8.28
10.70
8.10
9.30
3.35
5.00
26.60
e
e
e
e
e
e
e
e
Organochlorlde
Method
Detection
limit (ug/L)
0.004
0.003
0.006
0.009
0.004
0.014
0.011
0.004
0.012
0.002
0.014
0.004
0.066
0.006
0.023
0.003
0.083
0.176
0.24
nd
nd
nd
0.065
nd
nd
nd
Pesticides and PCBs.
Support Laboratory, Cincinnati, Ohio 45268.
e • Multiple peak response.
nd * not determined.



                                   8080 - 2
                                                          Revision
                                                          Date  September 1986

-------
TABLE 2.  DETERMINATION OF PRACTICAL QUANTITATION LIMITS (PQL) FOR VARIOUS
          MATRICES8


    Matrix                                                    Factor*
Ground water                                                     10
Low-level soil by sonlcatlon with GPC cleanup                   670
High-level soil and sludges by sonlcatlon                    10,000
Non-water mlsdble waste                                    100,000


     aSample  PQLs are highly  matrix-dependent.    The  PQLs listed herein are
     provided  for guidance and may not always be achievable.

     bPQL -  [Method detection limit  (Table 1)] X [Factor  (Table 2)].  For non-
     aqueous  samples, the factor 1s on a wet-weight basis.
                                   8080 - 3
                                                          Revision
                                                          Date  September 1986

-------
4.0  APPARATUS AND MATERIALS
     4.1  Gas chromatograph:
          4.1.1  Gas  Chromatograph:    Analytical  system  complete  with gas
     chromatograph  suitable  for   on-column   Injections  and  all  required
     accessories, Including detectors,  column  supplies, recorder, gases, and
     syringes.  A data system for  measuring peak heights and/or peak areas Is
     recommended.
          4.1.2  Columns:
               4.1.2.1   Column  1:  Supelcoport  (100/120 mesh) coated with 1.5%
          SP-2250/1.95%  SP-2401 packed  1n  a  1.8-m  x 4-mm I.D. glass column or
          equivalent.
               4.1.2.2   Column  2:  Supelcoport   (100/120  mesh) coated with 3%
          OV-1 1n  a 1.8-m x 4-mm  I.D. glass  column or  equivalent.
          4.1.3  Detectors:   Electron capture   (ECD) or halogen specific  (HSD)
      (I.e.,  electrolytic conductivity detector).
      4.2 Kuderna-Danlsh (K-D)  apparatus;
          4.2.1  Concentrator tube:   10-mL,  graduated  (Kontes K-570050-1025 or
      equivalent).   Ground-glass  stopper   1s used  to prevent evaporation of
      extracts
          4.2.2  Evaporation    flask:      500-mL    (Kontes   K-570001-500  or
      equivalent).   Attach to  concentrator tube  with  springs,
          4.2.3  Snyder column:    Three-ball   macro   (Kontes K-503000-0121 or
      equivalent).
          4.2.4  Snyder  column:     Two-ball micro   (Kontes  K-569001-0219 or
      equivalent).
      4.3  BolHng  chips;  Solvent extracted, approximately  10/40  mesh  (silicon
 carbide or  equivalent).
      4.4 Water   bath;     Heated,   with  concentric   ring   cover,   capable  of
 temperature control (+5*C).  The  bath should be used  1n  a hood.
      4.5  Volumetric flasks;   10-, 50-, and  100-mL,  ground-glass  stopper.
      4.6  M1crosyr1nge;   10-uL.
      4.7  Syringe:  5-mL.
      4.8  Vials;  Glass,  2-,  10-,  and  20-mL capacity with Teflon-Hned screw
 cap.
                                   8080 - 4
                                                          Revision
                                                          Date  September 1986

-------
5.0  REAGENTS

     5.1  Solvents:   Hexane,  acetone,  toluene,  Isooctane {2,2,4-trimethyl-
pentane) (pesticide quality or equivalent).

     5.2  Stock standard solutions;

          5.2.1  Prepare stock standard solutions at a concentration of
     1.00 ug/uL  by  dissolving  0.0100  g  of  assayed  reference material  in
     Isooctane and diluting to volume  1n  a  10-mL volumetric flask.  A small
     volume of toluene may be  necessary  to  put some pesticides 1n solution.
     Larger volumes can be  used  at   the  convenience  of  the analyst.  When
     compound purity 1s assayed to be  96%  or greater, the weight can be used
     without correction to calculate the  concentration of the stock standard.
     Commercially  prepared stock standards can be used at any concentration 1f
     they are certified by the manufacturer or by an Independent source.

          5.2.2  Transfer  the  stock  standard  solutions  Into Teflon-sealed
     screw-cap bottles.  Store at 4*C  and protect from light.  Stock standards
     should be checked  frequently  for  signs   of degradation or evaporation,
     especially just prior to preparing calibration standards from them.

          5.2.3  Stock  standard solutions must be  replaced after one year, or
     sooner  if comparison with check standards Indicates a problem.

     5.3  Calibration  standards:   Calibration  standards  at a minimum of five
 concentration  levelsTOTeach  parameter  of   Interest  are prepared through
 dilution  of  the  stock   standards   with  Isooctane.    One of the concentration
 levels should  be at  a  concentration  near,  but  above, the method detection
 limit. The  remaining  concentration  levels  should correspond to the expected
 range  of  concentrations found  in   real  samples or should define the working
 range  of  the GC.   Calibration solutions  must be replaced after six months, or
 sooner, 1f comparison  with check standards  Indicates a problem.

      5.4   Internal standards (if internal   standard  calibration is  used);  To
 use this  approach, the  analyst must  select  one or more internal standards that
 are similar  in analytical  behavior to  the  compounds of  interest.  The analyst
 must further demonstrate  that  the  measurement  of the  Internal standard  1s  not
 affected  by  method or  matrix Interferences.   Because of these limitations, no
 internal  standard  can  be  suggested that  1s  applicable  to all samples.

           5.4.1   Prepare   calibration    standards   at   a   minimum  of five
      concentration  levels   for  each  analyte   of  interest  as  described in
      Paragraph 5.3.

           5.4.2  To each  calibration  standard, add  a  known constant amount of
      one or more Internal  standards,  and dilute  to volume with Isooctane.

           5.4.3  Analyze  each  calibration  standard according to Section  7.0.
                                   8080 - 5
                                                          Revision      0
                                                          Date  September 1986

-------
     5.5  Surrogate standards;  The analyst  should monitor the performance of
the extraction, cleanup(when  used),  and  analytical  .system and the effec-
tiveness of the method  1n  dealing  with  each  sample matrix by spiking each
sample, standard, and reagent water  blank with pesticide surrogates.  Becau'se
GC/ECD data are much  more  subject  to  Interference  than GC/MS, a secondary
surrogate 1s to  be  used  when  sample  Interference  is  apparent.  Dibutyl-
chlorendate (DBC) 1s also subject  to  add  and base degradation.  Therefore,
two surrogate standards are added  to  each  sample; however, only one need be
calculated for recovery.   DBC  1s  the  primary  surrogate and should be used
whenever possible.  However,  1f  DBC  recovery  1s low or compounds interfere
with DBC, then  the  2,4,5,6-tetrachloro-meta-xylene  should  be evaluated for
acceptance.  Proceed with corrective  action  when  both surrogates are out of
limits for a sample  (Section  8.3).   Method 3500, Section 5.3.2,  Indicates the
proper procedure for preparing these surrogates.


6.0  SAMPLE COLLECTION, PRESERVATION, AND  HANDLING

     6.1  See  the  Introductory  material   to  this  chapter,  Organic Analytes,
Section 4.1.   Extracts  must  be  stored under  refrigeration and  analyzed within
40 days of  extraction.


7.0   PROCEDURE

      7.1   Extraction:

           7.1.1   Refer to Chapter Two for guidance on  choosing  the appropriate
      extraction  procedure.    In  general,   water  samples   are   extracted at a
      neutral,  or as 1s, pH  with   methylene chloride,  using  either Method 3510
      or 3520.   Solid samples are  extracted using either Method  3540 or 3550.

           7.1.2   Prior to gas chromatographlc analysis,  the  extraction solvent
      must be exchanged to hexane.    The  exchange Is  performed during the K-D
      procedures  listed 1n all  of  the  extraction  methods.    The exchange  is
      performed as follows.

                7.1.2.1  Following K-D of the methylene chloride extract to
           1 ml using the macro-Snyder column,   allow the apparatus to  cool  and
           drain  for at least 10 min.

                7.1.2.2  Increase the  temperature  of   the  hot  water bath  to
           about 90*C.  Momentarily  remove  the  Snyder  column,   add  50 mL  of
           hexane, a new boiling  chip,   and  reattach  the  macro-Snyder column.
           Concentrate the extract using 1  mL  of  hexane  to prewet the Snyder
           column.  Place the  K-D  apparatus  on  the   water  bath so  that the
           concentrator tube 1s partially  immersed  in  the  hot water.  Adjust
           the vertical position of the apparatus and the water temperature,  as
           required, to complete concentration in 5-10  m1n.  At the proper rate
           of distillation the balls of  the  column will actively chatter, but
           the chambers will not  flood.    When  the apparent volume of liquid
           reaches 1 mL, remove the  K-D  apparatus  and  allow it to drain and
           cool for at  least  10 m1n.

                                   8080 - 6
                                                          Revision      0	
                                                          Date  September 1986

-------
              7.1.2.3  Remove the Snyder column  and  rinse the flask and its
         lower joint Into the concentrator tube with 1-2 ml of hexane.  A
         5-mL syringe 1s recommended for  this operation.  Adjust the'extract
         volume  to  10.0  mL.    Stopper  the  concentrator  tube  and store
         refrigerated at 
-------
     7.4.3  Examples of GC/ECO  chromatograms  for various pesticides and
PCBs are shown in Figures 1 through 5.

     7.4.4  Prime the column as per Paragraph 7.3.2.

     7.4.5  DOT and endrfn are easily  degraded  In the Injection port ff
the Injection port or front of the  column  1s dirty.  This 1s the result
of buildup of high  boiling  residue  from  sample  injection.  Check for
degradation problems by  injecting  a  mid-level standard containing only
4,4'-DDT and endrin.    Look  for  the  degradation  products of 4,4'-DDT
(4,4'-DDE and 4,4'-DOD) and  endrin   (endrin ketone and endrin aldehyde).
If degradation of  either  DDT  or  endrin  exceeds  20%, take corrective
action before proceeding  with  calibration,  by  following the GC system
maintenance outlined 1n Section  7.7  of  Method 8000,  Calculate percent
breakdown as follows:

% breakdown     Total DDT degradation peak area (DDE + ODD)   1nn
for 4,4'-DDT       Total DDT peak area (DDT + DDE + ODD]    A iuu

X breakdown
for Endrin

Total endrin degradation peak area (endrin aldehyde + endrin ketone)  .Qn
 Total endrin peak ^rea  (endrin + endrin aldehyde + endrin ketone)

     7.4.6  Record the   sample  volume  Injected  and  the resulting peak
sizes  (In area units or  peak heights).

     7.4.7  Using either the   Internal  or  external calibration procedure
 (Method  8000), determine the  Identity and quantity  of each component peak
 in  the sample chromatogram  which  corresponds  to  the compounds  used for
calibration purposes.

     7.4.8   If peak detection  and   Identification  are   prevented  due to
 Interferences, the  hexane  extract   may  need  to  undergo cleanup using
 Method 3620.  The resultant extract(s)  may be  analyzed  by GC directly or
 may undergo  further cleanup to remove Sulfur using  Method 3660.

 7.5 Cleanup;

     7.5.1   Proceed with Method  3620,  followed  by,  if  necessary,  Method
 3660,  using  the  10-mL hexane  extracts obtained  from Paragraph  7.1.2.3.

      7.5.2   Following cleanup, the extracts should   be  analyzed  by  GC, as
 described 1n the previous  paragraphs and  1n Method  8000.

 7.6  Calculations  (exerpted  from U.S. FDA,  PAM):

      7.6.1   Calculation of Certain Residues:  Residues  which  are mixtures
 of two or more  components  present problems  in measurement.   When they are
 found  together,  e.g.,   toxaphene  and  DDT,  the   problem of quantitatlon
 becomes  even more difficult.    In  the  following  sections suggestions are
 offered  for handling  toxaphene,  chlordane,   PCB,   DDT,  and BHC.   A  column
 10% DC-200  stationary  phase   was  used  to  obtain  the chromatograms  in
 Figures  6-9.
                              8080 -  8
                                                     Revision     0
                                                     Date  September 1986

-------
      Column: 1.5% 9-2250*
              1JS% SP-2401 en Suptlcopon
      Ttmptrtiurt: 200° C
      Dtuctor: Eltctron Ciotun
0         4          I         12
             RETENTION TIME (MINUTES)
      Figure 1. Gas chromatogram of pesticides.
              8080  -  9
                                      Revision       0
                                      Date  September 1986

-------
        Column: 1.5% SP-2250-
               1.95\ S? 2401 on Swpticopon
        Ttmpmturt 20C°C
        Dructor: EUctron ClCtur*

           >	I	t
0         4          8         12
         RETENTION TIME (MINUTES)
    Figure 2. Gas chromatogr»m of chlordane.
         8080 - 10
                                Revision       0	
                                Date  September 1986

-------
                   Column; 1.S* 9-2250*
                          1J5N $P-2«01 en Sueticooon
                   TtfT»ptr»twrt. 200°C
                   Dmctor: Electron CiDtur*
          10        14       18
       HETENTION TIME (MINUTES)
22
26
Figure 3. Gas chromatogram of toxaphene.
         8080 -  11
                                 Revision      0
                                 Date   September 1986

-------
  Column :1.5% SJ-2250*
         US* SP 2401 on Sopticooon
  Ttmotr»turt: 200° C
  Otttctor: Electron CftDturt
I      •
           6          10          U
             RETENTION TIME (MINUTES)
22
      Figure 4. Gas chromatogram of PCS-1254.
                 8080  -  12
                                         Revision       0
                                         Date   September 1986

-------
Column: 1.S\SP2250*
       MS* SP-24C1 on Suwicooon
Ttmp«r«turt; 200eC
Octtcur: Eltctron Caeturt
                                  t    i
               10        U        18        22
             RETENTION TIME (MINUTES)
26
      Figure 5. G«s chromatogrvn of PCB-1260.
               8080  - 13
                                      Revision       0
                                      Date  September  1986

-------
    J..L
Fig. g—Basellne construction tor some typical gas chromatogrtphlc peaks.
a, symmetrical separated flat baseline; b and c, overlapping flat baseline;
d, separated (pen does not return to baseline between peaks); e, separated
sloping baseline; f,  separated (pen goes below baseline between peaks);
i, «- andy.BHC sloping baseline; h.•-,£-. and Y-BHC sloping baseline;
i, chlordane  flat baseline;  ), heptachlor and heptachlor epoxide super-
Imposed  on chlordane;  k, chair-shaped peaks, unsymmetrical peak;  1,
p,p'-DOT superimposed on toxapnene.
                          8080 -  14
                                                        Revision         o
                                                        Date   September 1986

-------
Fig. 7a—Baseline construction for multiple residues with standard
                         toxmphene.

   Pig. 7b~B*»ehne construction for multiple residues with coxa.
             phene, DOE and o.p'-, and p.p'-DOT.
                     8080  -  15
                                                   Revision        0
                                                   Date   September 1986

-------
 Fig. 6a-_Baseline construction for multiple residues: standard toxaphene.
Fig. 8fc— Baseline connructlon for multiple residues: rice bran with BHC,
                  toxephene, DOT, and meihoxychlor.
                         8080  - 16
                                                       Revision       0
                                                       Date   September  1986

-------
            Fig. 9»—Baseline construction for multiple residues:  standard chlordtne.
Fig. 9b—Baseline construolcD for multiple residues: rice bran with chlordane, toxapbetw, and DOT.
                                   8080  - 17
                                                                 Revision        0
                                                                 Date   September  1986

-------
     7.6.2  Toxaphene:  Quantitative calculation of toxaphene or Strobane
is difficult, but  reasonable  accuracy  can  be  obtained.  To calculate
toxaphene on GC/ECD:   (a)  adjust  sample  size  so that toxaphene major
peaks are 10-30%  full-scale  deflection  (FSD);  (b)  inject a toxaphene
standard that is  estimated  to  be  within  +10  ng  of  the sample; (c)
construct the baseline of standard  toxaphene between it extremities; and
(d) construct the baseline under  the  sample, using the distances of the
peak troughs to baseline on the  standard  as  a guide (Figures 7, 8, and
9).  This procedure   is  made  difficult  by  the  fact that the relative
heights and widths  of  the  peaks  in  the  sample  will probably not be
identical to the standard.    A  toxaphene  standard that has been passed
through a Flqrisil column will show  a  shorter retention time for peak X
and an enlargement of peak Y.

     7.6.3   Toxaphene and DDT:   If  DDT  is present, it will superimpose
itself on toxaphene peak  V.  To determine the approximate  baseline of the
DDT, draw a  line connecting  the trough  of  peaks U  and V  with the trough
of peaks W and  X and  construct  another line parallel to  this line  which
will just cut the  top of  peak  W   (Figure 61).  This procedure was tested
with ratios  of  standard toxaphene-DDT  mixtures  from 1:10  to 2:1 and the
results of added and  calculated DDT  and toxaphene by the  "parallel lines-
method of baseline construction were  within  10% of the  actual values in
all  cases.

           7.6.3.1   A  series  of  toxaphene   residues  have  been calculated
     using  total  peak area  for comparison   to the  standard and also  using
     area  of the  last four  peaks   only   1n  both sample  and standard. The
     agreement  between  the  results obtained  by the  two methods justifies
     the  use of the latter   method  for  calculating  toxaphene  in  a  sample
     where  the  early   eluting  portion   of   the toxaphene  chromatogram is
      interfered with  by other  substances.

           7.6.3.2   The    baseline    for   methoxychlor    superimposed  on
      toxaphene  (Figure  8b)  was constructed  by overlaying  the samples on  a
      toxaphene  standard of   approximately   the  same  concentration (Figure
      8a)  and viewing  the  charts  against  a  lighted  background.

      7.6.4   Chlordane  1s  a  technical  mixture   of at   least   11  major
 components   and  30  or  more    minor   ones.     Gas chromatography-mass
 spectrometry and nuclear  magnetic  resonance  analytical  techniques have
 been applied to the elucidation   of  the   chemical  structures  of  the many
 chlordane constituents.   Figure  9a  is   a  chromatogram  of standard  chlor-
 dane.   Peaks E  and F  are  responses to  trans-  and  cis-chlordane,  respec-
 tively.   These  are the   two  major components of  technical  chlordane, but
 the exact percentage of each in   the technical  material  is not  completely
 defined and is  not consistent  from  batch  to batch.  Other labelled  peaks
 in Figure 9a are  thought  to  represent:     A,  monochlorinated  adduct of
 pentachlorocyclopentadiene    with   cyclopentadiene;    B,    coelution  of
 heptachlor and  o-chlordene; C,   coelution  of /J-chlordene and 7-chlordene;
                              8080 - 18
                                                     Revision      0
                                                     Date  Seotember 1986

-------
D, a chlordane analog; G, coelution  of cis-nonachlor and "Compound K," a
chlordane Isomer.  The right "shoulder"  of  peak  F 1s  caused by trans-
nonachlor.

          7.6.4.1  The GC  pattern  of  a  chlordane  residue  may differ
     considerably from that of the  technical standard.  Depending on the
     sample substrate and Its history,  residues of chlordane can consist
     of almost  any  combination  of:    constituents  from the technical
     chlordane;  plant  and/or  animal   metabol1t1es;  and  products  of
     degradation caused  by  exposure  to  environmental  factors such as
     water and sunlight.  Only  limited information 1s available on which
     residue GC patterns are likely to  occur in which samples types, and
     even this Information may not be applicable to a situation where the
     route of exposure is unusual.  For example, fish exposed to a recent
     spill of  technical  chlordane  will  contain  a residue drastically
     different from a  fish  whose  chlordane  residue was accumulated by
     1ngest1on of  smaller  fish  or  of  vegetation,  which  1n turn had
     accumulated  residues  because  chlordane  was  in  the  water  from
     agricultural runoff.

          7.6.4.2  Because  of  this  Inability  to  predict  a chlordane
     residue GC  pattern, 1t is not  possible to prescribe a single method
     for  the quantltatlon of chlordane  residues.  The analyst must judge
     whether or  not the  residue's  GC  pattern  1s sufficiently similar to
     that of a technical chlordane   reference  material  to use the latter
     as a reference standard for quantltatlon.

          7.6.4.3  When  the chlordane  residue does not  resemble technical
     chlordane,   but   instead   consists   primarily   of    individual,
     identifiable peaks,  quantltate   each   peak  separately   against  the
     appropriate reference materials  and   report the  Individual  residues.
      (Reference  materials  are   available   for at   least   11   chlordane
     constituents, metabolites or degradation products  which  may  occur 1n
     the  residue.)

          7.6.4.4  When  the GC pattern  of   the  residue  resembles  that  of
     technical chlordane, quantitate  chlordane   residues by  comparing  the
     total  area  of   the  chlordane   chromatogram  from   peaks  A  through  F
      (Figure  9a) 1n   the  sample   versus   the   same   part of  the  standard
     chromatogram.   Peak G may be  obscured  in a  sample  by the  presence  of
     other   pesticides.     If  G   1s   not   obscured,   Include   it  In  the
     measurement for  both  standard  and  sample.   If the  heptachlor epoxide
     peak 1s  relatively  small,  include  it   as part of the total  chlordane
     area  for  calculation   of   the   residue.    If  heptachlor and/or
     heptachlor  epoxide  are  much   out  of   proportion   as   in  Figure  6j,
     calculate these  separately and  subtract their  areas from total area
     to  give  a corrected chlordane   area.    (Note that  octachlor  epoxide,
     metabolite  of   chlordane,  can   easily be  mistaken  for heptachlor
     epoxide  on  a  nonpolar  GC  column.)
                              8080 - 19
                                                     Revision      0
                                                     Date  September  1986

-------
         7.6.4.5  T6  measure   the   total   area   of   the  chlordane
    chromatogram, proceed as 1n Section  7.6.2  on toxaphene.  Inject an
    amount  of  technical  chlordane   standard  which  will  produce  a
    chromatogram 1n which peaks E and  F are approximately the same size
    as those  1n  the  sample  chromatograms.    Construct  the baseline
    beneath the standard from the beginning of peak A to the end of peak
    F as shown 1n Figure 9a.    Use the distance from the trough between
    peaks E and F to the baseline 1n the chromatogram of the standard to
    construct the baseline In the chromatogram of the sample.  Figure 9b
    shows how  the  presence  of  toxaphene  causes  the  baseline under
    chlordane to take an upward angle.   When  the size of peaks E and F
    1n standard and sample chromatograms are the same, the distance from
    the trough  of  the  peaks  to  the  baselines  should  be the same.
    Measurement of chlordane area should   be  done by total peak area 1f
    possible.
         NOTE:  A  comparison  has  been   made  of  the  total peak area
         Integration method and the addition  of peak heights method for
         several samples containing chlordane.    The peak heights A, B,
         C, D, E, and F  were  measured  1n millimeters from peak maximum
         of each to the  baseline  constructed under the total chlordane
         area  and were then added  together.   These results obtained by
         the two techniques are too close  to  Ignore this method of  "peak
         height addition"  as   a  means   of calculating  chlordane.  The
         technique  has  Inherent difficulties  because  not  all the  peaks
         are  symmetrical and  not all   are  present  1n the  same ratio  In
          standard  and  1n sample.     This  method  does   offer  a means  of
          calculating  results  1f  no   means of  measuring   total  area  1s
         practical.

     7.6.5   Polychlortnated blphenyls (PCBs):    Quantltatlon  of  residues
of PCB Involves problems similar to those encountered  In  the quantltatlon
of toxaphene,  Strobane,  and  chlordane:     1n   each  case,  the  chemical  1s
made up of numerous  compounds  and  so the chromatograms are  multi-peak;
also 1n each case the chromatogram  of  the residue  may not  match  that  of
the standard.

          7.6.5.1  Mixtures of PCB of various  chlorine contents were sold
     for many years 1n the U.S.  by  the Monsanto  Co.  under  the tradename
     Aroclor (1200 series and 1016).   Though these Aroclors  are no longer
     marketed, the PCBs remain 1n  the environment and are sometime found
     as residues In foods,  especially fish.

          7.6.5.2  PCB residues are quantltated  by  comparison to one or
     more of  the  Aroclor  materials,  depending   on  the chromatographlc
     pattern of the residue.  A  choice  must be made as to which Aroclor
     or mixture of Aroclors will  produce  a chromatogram most similar to
     that of the residue.   This  may  also Involve a judgment about what
     proportion of  the  different  Aroclors  to  combine  to produce the
     appropriate reference material.
                             8080 - 20
                                                    Revision      0
                                                    Date  September 1986

-------
              7.6.5.3  Quantltate PCB  residues  by  comparing  total area or
         height of residue  peaks  to  total  area  of  height  of peaks from
         appropriate Aroclor(s) reference materials.    Measure total area or
         height response from  common  baseline  under  all  peaks.  Use only
         those peaks from sample  that  can be attributed to chloroblphenyls.
         These peaks  must  also  be  present  1n  chromatogram  of reference
         materials.  Mixture  of  Aroclors  may  be  required to provide best
         match of GC patterns of sample and reference.

         7.6.6  DDT:  DDT found 1n  samples  often consists of both o.p1- and
    p,p'-DDT.  Residues of DDE  and  TDE  are  also frequently present.  Each
    Isomer of DDT and Its  metabolites  should  be quantHated using the pure
    standard of that compound and reported as such.

         7.6.7  Hexachlorocyclohexane  (BHC,  from  the  former name, benzene
    hexachlorlde):  Technical grade  BHC  is  a cream-colored amorphous solid
    with a very characteristic musty  odor;  1t  consists of a mixture of six
    chemically distinct isomers and  one or more heptachloro-cyclohexanes and
    octachloro-cyclohexanes.

              7.6.7.1  Commercial BHC preparations  may  show a wide variance
         in the percentage of  Individual  Isomers  present.  The elimination
         rate of the Isomers fed to rats  was  3 weeks for the a-, 7-, and 5-
         isomers and 14 weeks for the  /Msomer.   Thus 1t may be possible to
         have any   combination  of  the  various  isomers  in  different food
         commodities.   BHC  found  in  dairy  products  usually  has a large
         percentage of /Msomer.

              7.6.7.2   Individual  Isomers  (a, /J, 7, and 5) were  Injected  into
         gas  chromatographs  equipped with flame  1on1zat1on,  mlcrocoulometric,
         and  electron  capture detectors.     Response   for  the  four  Isomers  1s
         very nearly the   same  whether   flame  ionlzatlon  or mlcrocoulometric
         GLC  1s   used.     The  a-,  7-,   and  5-1somers   show   equal  electron
         affinity.  /J-BHC  shows a   much   weaker  electron  affinity  compared  to
         the  others Isomers.

               7.6.7.3   Quantltate  each  Isomer   (a,  /J,  7,   and  6}  separately
          against  a standard  of the respective   pure  Isomer,  using  a  GC  column
         which  separates  all  the  Isomers  from one  another.


8.0  QUALITY CONTROL

     8.1  Refer  to  Chapter  One   for  specific  quality   control  procedures.
Quality control  to validate sample  extraction is covered  1n Method  3500  and  In
the extraction method utilized.   If  extract cleanup  was  performed,  follow the
QC in Method 3600 and in the specific  cleanup method.

     8.2  Mandatory quality control   to  evaluate  the  GC system operation  1s
found in Method 8000,  Section 8.6.
                                  8080 - 21
                                                         Revision      0
                                                         Date  September 1986

-------
         8.2.1   The  quality  control  check   sample   concentrate  (Method 8000,
    Section  8.6)  should  contain  each   single-component  parameter of  Interest
    at  the following concentrations  1n  acetone:    4,4'-DDD,  10 ug/mL; 4,4'-
    DDT, 10  ug/mL; endosulfan   II,   10   ug/mL;   endosulfan  sulfate, 10 ug/mL;
    endrln,  lOug/mL; and any other  single-component  pesticide, 2  ug/mL.   If
    this method 1s   only  to  be  used   to   analyze  for  PCBs,  chlordane,  or
    toxaphene,  the   QC  check   sample   concentrate   should   contain  the most
    representative millt1-component parameter at  a   concentration of  50 ug/mL
    1n  acetone.

         8.2.2  Table 3  Indicates  the   calibration  and QC acceptance  criteria
    for this  method.    Table  4   gives method accuracy  and  precision  as
    functions of concentration for the  analytes of  Interest.  The contents  of
    both Tables should be used to   evaluate a laboratory's  ability  to perform
    and generate acceptable data by this method.

    8.3  Calculate surrogate standard   recovery  on  all samples, blanks, and
spikes.   Determine 1f  the  recovery  1s  within limits (limits established  by
performing  QC procedures outlined 1n Method  8000, Section 8.10).

          8.3.1  If recovery 1s not within limits, the following  1s  required.

                •  Check to  be  sure  there   are  no  errors  1n  calculations,
                  surrogate solutions  and  Internal  standards.   Also,  check
                  Instrument performance.

                •  Recalculate the data and/or reanalyze  the extract 1f  any  of
                  the  above checks reveal a  problem.

                •  Reextract and reanalyze the sample  1f none of the above are
                  a problem or flag the data as  "estimated concentration."

     8.4  GC/MS confirmation;  Any compounds confirmed by two columns may also
be confirmed by GC/MST7the  concentration  Is  sufficient for detection  by
GC/MS as determined  by the laboratory generated  detection limits.

          8.4.1   The  GC/MS would normally  require  a minimum concentration of
     10 ng/uL 1n  the  final extract, for each single-component compound.

          8.4.2  The pesticide extract  and associated blank  should  be analyzed
     by GC/MS as  per Section 7.0 of Method 8270.

          8.4.3  The confirmation  may  be   from the  GC/MS analysis  of  the
     base/neutral-add extractables extracts  (sample   and blank).   However,  1f
     the compounds are  not  detected   1n  the   base/neutral-add extract even
     though   the concentration  1s  high  enough,   a  GC/MS analysis  of  the
     pesticide  extract should  be performed.

          8.4.4  A reference standard of the compound  must  also  be  analyzed by
     GC/MS.   The concentration of   the   reference  standard must be at a  level
     that  would demonstrate   the  ability   to  confirm  the  pest1c1des/PCBs
     Identified by GC/ECD.


                                   8080  -  22
                                                          Revision      0
                                                          Date  September  1986

-------
9.0  METHOD PERFORMANCE

     9.1  The method  was  tested  by  20  laboratories  using  reagent water,
drinking water, surface water, and  three Industrial wastewaters spiked at six
concentrations.  Concentrations used 1n the  study  ranged from 0.5 to 30 ug/L
for single-component pesticides and from  8.5  to 400 ug/L for multl-component
parameters.  Single operator precision, overall precision, and method accuracy
were found to be directly  related  to  the concentration of the parameter and
essentially Independent of the  sample  matrix.   Linear equations to describe
these relationships for a flame 1on1zat1on detector are presented 1n Table 4,

     9.2  The accuracy and precision obtained will be determined by the sample
matrix,  sample-preparation  technique,   optional   cleanup  techniques,  and
calibration procedures used.


10.0  REFERENCES

1.  U.S. EPA,  "Development  and  Application  of  Test Procedures for Specific
Organic Toxic Substances  1n  Wastewaters,  Category  10: Pesticides and PCBs,"
Report for EPA  Contract 68-03-2605.

2.  U.S. EPA,  "Interim  Methods  for   the  Sampling  and  Analysis of Priority
Pollutants 1n  Sediments and F1sh Tissue," Environmental Monitoring and Support
Laboratory, Cincinnati, OH 45268, October 1980.

3.  Pressley,  T.A., and   J.E.   Longbottom,   "The  Determination of Organohallde
Pesticides and  PCBs 1n  Industrial   and Hunlcipal  Wastewater: Method 617," U.S.
EPA/EMSL, Cincinnati, OH,  EPA-600/4-84-006,  1982.

4.  "Determination of  Pesticides  and  PCB's   in   Industrial   and  Municipal
Wastewaters, U.S.  Environmental  Protection  Agency,"  Environmental Monitoring
and Support Laboratory, Cincinnati, OH 45268,  EPA-600/4-82-023,  June 1982.

5.  Goerlltz,  D.F. and  L.M. Law,  Bulletin for Environmental Contamination  and
Toxicology, 6,  9,  1971.

6.  Burke,  J.A.,   "Gas  Chromatography  for   Pesticide  Residue  Analysis; Some
Practical   Aspects,"   Journal   of   the  Association   of   Official  Analytical
Chemists,  48,  1037,  1955.

7.  Webb,  R.G.   and  A.C.  McCall,   "Quantitative  PCB Standards  for  Electron
Capture  Gas   Chromatography,"   Journal   of   Chromatographlc  Science,  11, 366,
 1973.

8.  Millar,  J.D.,  R.E.  Thomas  and   H.J.   Schattenberg,   "EPA  Method Study  18,
Method 608:   Organochlorine  Pesticides   and  PCBs,"   U.S.   EPA/EMSL,  Research
 Triangle Park,  NC, EPA-600/4-84-061,  1984.

 9.   U.S.  EPA 40 CFR Part  136,  "Guidelines  Establishing Test Procedures for  the
 Analysis of Pollutants Under the Clean Water Act; Final  Rule and Interim Final
 Rule  and Proposed Rule,"  October 26,  1984.
                                   8080 - 23
                                                          Revision      0
                                                          Date  September 1986

-------
10.  Provost, L.P. and R.S. Elder,  "Interpretation of Percent Recovery Data,"
American Laboratory, lj>, pp. 58-63, 1983.

11.  U.S. Food and Drug  Administration,  Pesticide Analytical Manual, Vol.  1,
Oune 1979.

12.  Sawyer, L.D., JAOAC,  56,  1015-1023   (1973), 61 272-281  (1978),  61 282-291
(1978).

13.  Official Methods of Analysis   of  the   Association  of Official  Analytical
Chemists, 12th  Edition; Changes   1n Methods,   JAOAC   61,  465-466  (1978),  Sec.
29.018.
                                   8080 - 24
                                                          Revision
                                                          Date   September  1986

-------
TABLE 3.  QC ACCEPTANCE CRITERIA*
Parameter
Aldrln
a-BHC
tf-BHC
ff-BHC
7-BHC
Chlordane
4,4'-DDD
4,4'-DDE
4,4'-DDT
Dleldrln
Endosulfan I
Endosulfan II
Endosulfan Sulfate
Endrln
Heptachlor
Heptachlor epoxlde
Toxaphene
PCB-1016
PCB-1221
PCB-1232
PCB-1242
PCB-1248
PCB-1254
PCB-1260
Test
cone.
(ug/L)
2.0
2.0
2.0
2.0
2.0
50
10
2.0
10
2.0
2.0
10
10
10
2.0
2.0
50
50
50
50
50
50
50
50
Limit
for s
(ug/L)
0.42
0.48
0.64
0.72
0.46
10.0
2.8
0.55
3.6
0.76
0.49
6.1
2.7
3.7
0.40
0.41
12.7
10.0
24.4
17.9
12.2
15.9
13.8
10.4
.tenge
for 7
(ug/L)
i. 08-2. 24
.98-2.44
U. 78-2. 60
1.01-2.37
0.86-2.32
27.6-54.3
4.8-12.6
1.08-2.60
4.6-13.7
1.15-2.49
1.14-2.82
2.2-17.1
3.8-13.2
5.1-12.6
0.86-2.00
1.13-2.63
27.8-55.6
30.5-51.5
22.1-75.2
14.0-98.5
24.8-69.6
29.0-70.2
22.2-57.9
18.7-54.9
Range
P, PS
(X)
42-122
37-134
17-147
19-140
32-127
45-119
31-141
30-145
25-160
36-146
45-153
D-202
26-144
30-147
34-111
37-142
41-126
50-114
15-178
10-215
39-150
38-158
29-131
8-127
      s » Standard deviation of four recovery measurements, 1n .ug/L.

      7 = Average recovery for four recovery measurements, 1n ug/L.

      P, Ps = Percent recovery measured.

      D = Detected; result must be greater than zero.

      aCr1ter1a from 40 CFR Part 136 for  Method 608.  These criteria are based
 directly upon the method performance  data  1n  Table 4.  Where necessary, the
 limits for recovery have been broadened  to assure applicability of the limits
 to concentrations below those used to develop Table 4.
                                   8080 - 25
                                                           Revision      0
                                                           Date  September  1986

-------
TABLE 4.  METHOD ACCURACY AND PRECISION AS FUNCTIONS OF CONCENTRATION3
Parameter
Aldrln
a-BHC
fl-BHC
6-BHC
7-BHC
Chlordane
4,4'-DDD
4,4'-DDE
4,4'-DOT
D1eldr1n
Endosulfan I
Endosulfan II
Endosulfan Sulfate
Endrln
Heptachlor
Heptachlor epoxlde
Toxaphene
PCB-1016
PCB-1221
PCB-1232
PCB-1242
PCB-1248
PCB-1254
PCB-1260
Accuracy, as
recovery, x'
(ug/L)
0.81C+0.04
0.84C+0.03
0.81C+0.07
0.81C+0.07
0.82C-0.05
0.82C-0.04
0.84C+0.30
0.85O0.14
0.93C-0.13
0.90C+0.02
0.97C+0.04
0.93C+0.34
0.89C-0.37
0.89C-0.04
0.69C+0.04
0.89C+0.10
0.80C+1.74
0.81C+0.50
0.96C+0.65
0.91C+10.79
0.93C+0.70
0.97C+1.06
0.76C+2.07
0.66C+3.76
Single analyst
precision, sr'
'(ug/L)
0.167-0.04
0.137+0.04
0.227+0.02
0.187+0.09
0.127+0.06
0.137+0.13
0.207-0.18
0.137+0.05
0.177+0.35
0.127+0.19
0.107+0.07
0.417-0.65
0.137+0.33
0.207+0.25
0.067+0.13
0.187-0.11
0.097+3.20
0.137+0.15
0.297-0.76
0.217-1.93
0.117+1.40
0.177+0.41
0.157+1.66
0.227-2.37
Overall
precision,
S' (ug/L)
0.207-0.01
0.237-0.00
0.337-0.95
0.257+0.03
0.227+0.04
0.187+0.18
0.277-0.14
0.28T-0.09
0.31x-0.21
0.167+0.16
0.187+0.08
0.477-0.20
0.247+0.35
0.247+0.25
0.167+0.08
0.257-0.08
0.207+0.22
0.157+0.45
0.357-0.62
0.317+3.50
0.217+1.52
0.257-0.37
0.177+3.62
0.397-4.86
      x1  = Expected  recovery  for  one  or  more  measurements  of  a  sample
            containing a concentration of C, in ug/L.

      sr' = Expected single analyst  standard  deviation  of measurements at an
            average concentration of 7, in ug/L.

      S1  = Expected interlaboratory standard  deviation  of measurements at an
            average concentration found of 7, in ug/L.

      C   = True value for the concentration, in ug/L.

      7   = Average recovery found for measurements of samples containing a
            concentration of C, in ug/L.
                                   8080 - 26
                                                          Revision      0
                                                          Date  September 1986

-------
                                          METHOD 8080

                               ooG«MOCHLO*!NE PESTICIDES c PCB>
C   --•    )
 7. i . 1
     O
        Cnoose
     eppropr iete
     extractIon
      procedure
  (cee Cnepter 2)
                                                      7.4
    Perform CC
  analysis (see
   Method 8000)
 7.1.8
        Exchange
        •xtr»ct-
  lon solvent to
        heiccne
    during micro
  K-D procedures
  7.Z
     Set g»«
  chrometogr«pny
  7.3
        Befer to
     Method 6OOO
     'or proper
     celloretlon
     teennlaues
 7.3.2
        pe»K
detection 6 iden-
   tification
     prevent-
       ed?
    Cleanup
 using Method
36ZO end 3660
if neeescery
                             Methods or
                            calculation of
                        toxepnene. cnioraane.
                        PCB. DOT.  and One ere
                           handled in this
                             • ret Ion
        Prieie or
      deectivete
      CC column
  prior to deily
    calibration
     O
                                      8080  - 27
                                                                 Revision       0
                                                                 Date   September 1986

-------
00
o
00
o

-------
                                 METHOD 8080A

           ORGANOCHLORINE  PESTICIDES AND  POLYCHLORINATED  BIPHENYLS
                            BY GAS CHROMATOGRAPHY
1.0   SCOPE AND APPLICATION

      1.1   Method  8080  is  used  to  determine the  concentration of  various
organochlorine pesticides and polychlorinated biphenyls (PCBs).  The following
compounds can be determined by this method:
      Compound Name
CAS No."
Aldrin
a-BHC
/3-BHC
5-BHC
7-BHC (Lindane)
Chlordane (technical)
4,4'-DDD
4,4'-DDE
4,4'-DDT
Dieldrin
Endosulfan I
Endosulfan II
Endosulfan sulfate
Endrin
Endrin aldehyde
Heptachlor
Heptachlor epoxide
4,4'-Methoxychlor
Toxaphene
Aroclor-1016
Aroclor-1221
Aroclor-1232
Aroclor-1242
Aroclor-1248
Aroclor-1254
Aroclor-1260
309-00-2
319-84-6
319-85-7
319-86-8
58-89-9
12789-03-6
72-54-8
72-55-9
50-29-3
60-57-1
959-98-8
33212-65-9
1031-07-8
72-20-8
7421-93-4
76-44-8
1024-57-3
72-43-5
8001-35-2
12674-11-2
1104-28-2
11141-16-5
53469-21-9
12672-29-6
11097-69-1
11096-82-5
a  Chemical Abstract Services Registry Number.

      1.2   Table  1  lists  the  method  detection  limit  for  each  compound  in
organic-free reagent water.  Table 2 lists the estimated quantitation limit (EQL)
for other matrices.
                                   8080A  -  1
                  Revision 1
              September 1994

-------
2.0   SUMMARY OF METHOD

      2.1   Method 8080 provides gas chromatographic conditions for the detection
of ppb concentrations of certain organochlorine pesticides and PCBs.  Prior to
the use of this method, appropriate sample extraction techniques must be used.
Both neat  and diluted organic  liquids (Method  3580,  Waste Dilution)  may be
analyzed by  direct injection.   A 2  to 5 ^L  sample  is  injected   into  a  gas
chromatograph (GC) using the  solvent  flush technique,  and compounds in the GC
effluent are detected by an electron capture  detector (ECD) or an electrolytic
conductivity detector  (HECD).

      2.2   The sensitivity of Method  8080 usually depends on the concentration
of  interferences  rather  than on  instrumental  limitations.   If interferences
prevent detection of the  analytes,  Method 8080  may also be performed on samples
that have undergone cleanup.  Method 3620,  Florisil Column  Cleanup, by itself or
followed by Method 3660,  Sulfur Cleanup, may be used to eliminate interferences
in the analysis.


3.0   INTERFERENCES

      3.1   Refer to Methods 3500, 3600, and  8000.

      3.2   Interferences  by phthalate esters  can  pose   a  major   problem in
pesticide  determinations when  using   the  electron  capture  detector.    These
compounds  generally  appear  in  the chromatogram  as  large  late-eluting peaks,
especially  in  the 15% and  50% fractions  from the  Florisil cleanup.   Common
flexible plastics contain varying  amounts  of phthalates.  These phthalates are
easily extracted  or  leached  from  such materials  during laboratory  operations.
Cross contamination of  clean glassware  routinely occurs when plastics  are handled
during extraction  steps, especially when  solvent-wetted surfaces are handled.
Interferences from phthalates can  best  be minimized by  avoiding  contact with any
plastic materials. Exhaustive cleanup  of reagents and  glassware may  be required
to  eliminate  background  phthalate  contamination.     The contamination  from
phthalate  esters  can  be  completely   eliminated with  a   microcoulometric  or
electrolytic conductivity detector.


4.0   APPARATUS AND MATERIALS

      4.1   Gas chromatograph

            4.1.1 Gas  Chromatograph:    Analytical   system  complete with  gas
      chromatograph  suitable   for on-column   injections  and  all  required
      accessories, including  detectors, column supplies,  recorder,  gases,  and
      syringes.  A data  system  for measuring  peak heights and/or peak areas is
      recommended.

            4.1.2 Columns

                  4.1.2.1     Column 1: Supelcoport  (100/120 mesh)  coated with
            1.5% SP-2250/1.95%  SP-2401 packed in  a 1.8 m x 4 mm ID glass column
            or equivalent.
                                   8080A -  2
    Revision 1
September 1994

-------
                  4.1.2.2     Column 2:   Supelcoport  (100/120 mesh) coated with
            3% OV-1 in a 1.8 m x 4 mm ID glass column or equivalent.

            4.1.3 Detectors:     Electron   capture   (ECD)   or   electrolytic
      conductivity detector (HECD).

      4.2   Kuderna-Danish (K-D) apparatus:

            4.2.1 Concentrator tube:   10  ml, graduated (Kontes K-570050-1025 or
      equivalent).  A  ground-glass stopper is used  to  prevent  evaporation of
      extracts.

            4.2.2 Evaporation   flask:      500   ml    (Kontes  K-570001-500   or
      equivalent).   Attach  to  concentrator  tube  with  springs,  clamps,  or
      equivalent.

            4.2.3 Snyder column:   Three ball  macro (Kontes K-503000-0121 or
      equivalent).

            4.2.4 Snyder  column:    Two   ball  micro  (Kontes  K-569001-0219 or
      equivalent).

            4.2.5 Springs -  1/2 inch (Kontes K-662750 or equivalent).

      4.3   Boiling chips:   Solvent extracted,  approximately 10/40 mesh (silicon
carbide or equivalent).

      4.4   Water  bath:     Heated,  with  concentric  ring  cover,  capable  of
temperature control (±5°C).   The bath should  be  used  in  a  hood.

      4.5   Volumetric flasks, Class A:   sizes as appropriate with ground-glass
stoppers.

      4.6   Microsyringe:  10 /nL.

      4.7   Syringe:  5 ml.

      4.8   Vials:  Glass, 2, 10, and 20 ml capacity with  Teflon-lined screw caps
or crimp tops.

      4.9   Balances:   Analytical,  0,0001 g and Top loading,  0.01 g.


5.0   REAGENTS

      5.1   Reagent grade chemicals shall be used in all  tests. Unless otherwise
indicated, it is intended that all  reagents shall conform to the specifications
of the Committee on Analytical  Reagents of  the American Chemical Society, where
such specifications are available. Other grades may be used, provided  it is first
ascertained that  the reagent  is  of sufficiently  high purity  to  permit its  use
without lessening the accuracy of the determination.
                                   8080A  -  3                         Revision 1
                                                                September 1994

-------
      5.2   Organic-free reagent water - All references to water in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.3   Solvents

            5.3.1 Hexane, C6H14  - Pesticide quality or equivalent.

            5.3.2 Acetone,  CH3COCH3 -  Pesticide quality  or  equivalent.

            5.3.3 Toluene, C6H5CH3 -  Pesticide quality or equivalent.

            5.3.4 Isooctane, (CH3)3CCH2CH(CH3)2 - Pesticide quality or equivalent.

      5.4   Stock standard solutions:

            5.4.1 Prepare  stock  standard solutions  at  a  concentration  of
      1000 mg/L  by  dissolving   0.0100  g  of  assayed  reference  material  in
      isooctane  and  diluting  to volume in a 10 ml volumetric  flask.   A small
      volume of  toluene  may be  necessary to put  some pesticides  in solution.
      Larger volumes can be  used  at  the convenience  of  the  analyst.   When
      compound purity is  assayed to be 96% or  greater,  the  weight can be used
      without correction to calculate the concentration  of the stock standard.
      Commercially prepared stock standards can be used at any concentration if
      they are certified by the manufacturer or by an independent source.

            5.4.2 Transfer the stock standard  solutions into vials with Teflon-
      lined screw  caps  or crimp tops.   Store at  4°C and protect  from light.
      Stock standards should be checked frequently for signs of degradation or
      evaporation, especially just prior to preparing  calibration standards from
      them.

            5.4.3 Stock standard solutions must be replaced after one year, or
      sooner if  comparison with check standards indicates a problem.

      5.5   Calibration standards:  Calibration standards at a minimum of five
concentrations for each parameter of interest are prepared through dilution of
the stock  standards  with  isooctane.   One  of  the  concentrations should be at a
concentration  near,  but  above, the  method  detection limit.   The remaining
concentrations should correspond to the expected range of concentrations found
in  real  samples  or  should  define the working range  of the GC.   Calibration
solutions must be replaced after six months,  or sooner, if comparison with check
standards  indicates  a problem.

      5.6   Internal standards  (if  internal  standard calibration is used):  To
use this approach, the analyst  must select one or more internal standards that
are similar in analytical behavior  to  the compounds  of  interest.   The analyst
must further demonstrate  that the measurement  of  the internal  standard is not
affected by method  or matrix  interferences.   Because of these limitations, no
internal standard can be  suggested  that is applicable to all samples.

            5.6.1 Prepare   calibration   standards  at   a  minimum   of  five
      concentrations for each analyte of  interest as described in Sec. 5.5.
                                   8080A -  4                         Revision 1
                                                                September 1994

-------
            5.6.2 To each calibration standard, add a known constant amount of
      one or more internal standards, and dilute to volume with isooctane.

            5.6.3 Analyze each calibration standard according to Sec. 7.0.

      5.7   Surrogate standards:   The analyst  should monitor the performance of
the extraction,  cleanup  (when used), and analytical  system and the effectiveness
of  the  method  in  dealing with  each  sample  matrix  by  spiking  each  sample,
standard,  and  organic-free  reagent  water  blank  with  pesticide  surrogates.
Because GC/ECD data are much more  subject to interference than GC/MS, a secondary
surrogate is  to be  used when sample  interference  is  apparent.   Two surrogate
standards (tetrachloro-m-xylene (TCMX) and decachlorobiphenyl)  are added to each
sample;  however,  only  one  need  be  calculated  for  recovery.    Proceed  with
corrective action when both surrogates are out  of limits for a sample (Sec. 8.3).
Method 3500 indicates the proper procedure for preparing these surrogates.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See the  introductory material  to this chapter, Organic Analytes, Sec.
4.1.  Extracts must  be stored under refrigeration and analyzed within 40 days of
extraction.
7.0   PROCEDURE

      7.1   Extraction:

            7.1.1 Refer to Chapter Two for guidance on choosing the appropriate
      extraction  procedure.    In  general,  water  samples  are  extracted  at  a
      neutral, or as is, pH with methylene chloride,  using either Method 3510 or
      3520.  Solid samples are extracted using Method 3540, 3541,  or 3550.

            7.1.2 Prior to gas chromatographic  analysis,  the extraction solvent
      must be  exchanged to  hexane.   The exchange is  performed during  the K-D
      procedures  listed  in all of  the extraction methods.   The  exchange is
      performed as follows.

                  7.1.2.1     Following K-D of  the methylene chloride extract to
            1 mL using the macro-Snyder column, allow  the apparatus to cool and
            drain for at least 10 min.

                  7.1.2.2     Increase the temperature of the hot water bath to
            about 90°C.   Momentarily  remove the  Snyder  column,  add 50  mL of
            hexane,  a new boiling chip,  and reattach  the macro-Snyder column.
            Concentrate the extract using  1  mL of hexane to prewet the Snyder
            column.    Place  the K-D  apparatus  on the water bath   so  that the
            concentrator tube  is  partially  immersed  in  the hot water.   Adjust
            the vertical  position  of the apparatus and the water temperature, as
            required, to complete  concentration in 5-10 min.  At the proper rate
            of distillation the balls of the column will actively  chatter, but
            the chambers will  not flood.   When the  apparent  volume  of liquid
                                   8080A  -  5                         Revision 1
                                                                September 1994

-------
            reaches 1 ml,  remove  the  K-D apparatus and allow  it  to drain and
            cool  for at least 10 min.

                  7.1.2.3     Remove the Snyder column and rinse the flask and
            its lower joint into the  concentrator  tube  with  1-2 mL of hexane.
            A  5 ml  syringe  is  recommended  for  this operation.    Adjust the
            extract volume to 10.0 ml.   Stopper the concentrator tube and store
            refrigerated at  4°C,  if further processing  will not  be performed
            immediately.  If the extract will be stored longer than  two days, it
            should be transferred  to  a vial with a Teflon-lined  screw  cap or
            crimp top.   Proceed  with  gas chromatographic  analysis if further
            cleanup is not required.

      7.2   Gas chromatography conditions (Recommended):

            7.2.1 Column 1:

            Carrier gas (5% methane/95% argon) flow rate:    60 mL/min
            Column temperature:                              200°C  isothermal

            When  analyzing  for  the  low molecular  weight  PCBs  (PCB 1221-PCB
      1248), it is advisable to  set the oven temperature to 160°C.

            7.2.2 Column 2:

            Carrier gas (5% methane/95% argon) flow rate:    60 mL/min
            Column temperature:                              200°C  isothermal

            When  analyzing  for  the  low molecular  weight  PCBs  (PCB 1221-PCB
      1248), it is advisable to  set the oven temperature to 140°C.

            7.2.3 When analyzing for most or all of the analytes  in  this method,
      adjust the  oven  temperature and  column  gas  flow to  provide sufficient
      resolution for accurate quantitation  of the analytes.  This will normally
      result in a retention time  of 10 to 12 minutes for 4,4'-DDT, depending on
      the packed column used.

      7.3   Calibration: Refer to Method 8000 for proper  calibration techniques.
Use Table 1  and especially Table 2 for guidance on selecting the  lowest point on
the calibration curve.

            7.3.1 The procedure  for  internal or  external  calibration  may be
      used.   Refer to Method 8000 for a description  of  each of these procedures.

            7.3.2 Because  of the  low  concentration  of  pesticide  standards
      injected on a GC/ECD, column adsorption may be a problem when the GC has
      not been used for  a  day.   Therefore, the GC column  should  be primed or
      deactivated by injecting a PCB or pesticide standard mixture approximately
      20 times more concentrated  than  the  mid-concentration  standard.   Inject
      this prior to beginning initial  or daily calibration.
                                   8080A  -  6                         Revision 1
                                                                September 1994

-------
7.4   Gas chromatographic analysis:

      7.4.1 Refer to Method 8000.   If  the internal  standard calibration
technique is used, add 10 /iL of internal standard to the sample prior to
injection.

      7.4.2 Method 8000 provides  instructions  on  the analysis sequence,
appropriate dilutions,  establishing daily  retention time  windows,  and
identification criteria.   Include  a mid-concentration standard after each
group of 10 samples in the analysis sequence.

      NOTE: A 72 hour sequence is not required with this method.

      7.4.3 Examples of GC/ECD chromatograms  for  various pesticides and
PCBs are shown in Figures 1 through 5.

      7.4.4 Prime the column as per Sec. 7.3.2.

      7.4.5 DDT and endrin are easily  degraded  in  the  injection  port if
the injection port or  front of the column is dirty.   This  is  the result of
buildup  of high  boiling  residue  from sample  injection.    Check  for
degradation problems by injecting  a mid-concentration standard containing
only 4,4'-DDT and endrin.  Look for the degradation products of 4,4'-DDT
(4,4'-DDE and 4,4'-DDD) and endrin  (endrin  ketone  and  endrin aldehyde).
If degradation of  either DDT or endrin exceeds  20%, take corrective action
before proceeding  with  calibration,  by following the  GC system maintenance
outlined in of Method 8000.  Calculate  percent breakdown as follows:

                          Total DDT degradation peak area (DDE + ODD)
      % breakdown       =	   x 100
      for 4,4'-DDT          Total  DDT peak area (DDT + DDE + ODD)

                          Total endrin  degradation peak area
                          (endrin aldehyde + endrin ketone)
      % breakdown       = 	   x 100
      for Endrin           Total  endrin peak area (endrin +
                           endrin aldehyde + endrin ketone)

      7.4.6 Record the sample volume injected and the resulting peak sizes
(in area units or peak heights).

      7.4.7 Using either  the  internal  or  external  calibration procedure
(Method 8000), determine  the identity and quantity of each component peak
in the  sample  chromatogram which  corresponds to the compounds  used for
calibration purposes.

      7.4.8 If peak  detection and  identification  are  prevented due  to
interferences, the hexane  extract may need to undergo cleanup using Method
3620.   The  resultant  extract(s)  may be analyzed  by GC  directly or may
undergo further cleanup to remove sulfur using Method 3660.
                            8080A  -  7                         Revision 1
                                                          September 1994

-------
                                           '0
7.5   Cleanup:

      7.5.1 Proceed with Method 3620,  followed  by,  if necessary, Method
3660, using the 10 ml hexane extracts obtained from Sec. 7.1.2.3.

      7.5.2 Following cleanup, the extracts should be analyzed by GC, as
described in the previous sections and in Method 8000.

      7.5.3 If only  PCBs  are to  be  measured  in a  sample,  the sulfuric
acid/permanganate  cleanup   (Method  3665),  followed  by Silica  Cleanup
(Method 3630) or Florisil  Cleanup (Method 3620), is recommended.

7.6   Calculations (excerpted from U.S. FDA, RAM):

      7.6.1 Calculation of Certain Residues:  Residues which are mixtures
of two or more components present  problems  in measurement.  When they are
found  together,  e.g., toxaphene  and  DDT,  the  problem  of  quantitation
becomes even more difficult.   In  the following  sections suggestions are
offered  for  handling toxaphene,  chlordane,  PCB,  DDT,  and  BHC.   A 10%
DC-200 stationary phase column  was used to obtain  the  chromatograms in
Figures 6-9.

      7.6.2 Toxaphene:  Quantitative  calculation of toxaphene or Strobane
is  difficult,  but reasonable  accuracy can be  obtained.   To  calculate
toxaphene on GC/ECD:   (a) adjust sample size so that toxaphene major peaks
are  10-30% full-scale deflection  (FSD);  (b)  inject  a toxaphene standard
that  is  estimated to be  within +10 ng of  the sample;  (c)  construct the
baseline of standard toxaphene between  its  extremities;  and (d) construct
the baseline under the sample,  using  the  distances of the peak troughs to
baseline on the standard as  a guide (Figures 7, 8,  and  9).  This procedure
is made difficult by the fact that the  relative heights  and widths of the
peaks  in  the  sample  will  probably not be  identical to  the  standard.   A
toxaphene standard  that has been  passed through  a  Florisil  column will
show a shorter retention time for peak X and an  enlargement of peak Y.

      7.6.3 Toxaphene and DDT:   If DDT  is  present,  it  will  superimpose
itself on toxaphene peak V.   To  determine the approximate baseline of the
DDT, draw a line  connecting  the trough of peaks U and V with the trough of
peaks W and X and construct  another line  parallel to this line which will
just cut  the top of  peak W  (Figure 61).   This procedure was tested with
ratios of standard toxaphene-DDT mixtures from 1:10 to  2:1 and the results
of added  and calculated DDT and toxaphene by the "parallel  lines" method
of  baseline  construction  were  within  10% of the  actual values  in all
cases.

            7.6.3.1     A   series  of  toxaphene   residues   have  been
      calculated using total peak  area  for  comparison to the standard and
      also using  area of the  last four peaks  only  in  both  sample and
      standard.  The agreement  between the results obtained  by the two
      methods  justifies  the  use  of  the  latter method  for  calculating
      toxaphene  in   a  sample where  the  early  eluting  portion  of the
      toxaphene chromatogram is interfered with  by other substances.
8080A - 8                         Revision  1
                             September  1994

-------
            7.6.3.2     The  baseline  for  methoxychlor  superimposed  on
      toxaphene (Figure 8b) was constructed by overlaying the samples on
      a toxaphene  standard of approximately the same concentration (Figure
      8a) and viewing the charts against a lighted background.

      7.6.4 Chlordane  is  a  technical  mixture  of  at  least  11  major
components  and  30  or  more   minor   ones.     Gas  chromatography-mass
spectrometry and  nuclear magnetic resonance  analytical  techniques  have
been applied to the  elucidation of the chemical  structures  of  the  many
chlordane constituents.  Figure 9a  is  a chromatogram of standard chlor-
dane.  Peaks E and F are responses  to  trans-  and cis-chlordane,  respec-
tively.  These are the two major  components  of technical chlordane,  but
the exact percentage of each in the technical  material is not completely
defined and is not consistent from batch to batch.  Other labelled peaks
in  Figure  9a are  thought  to represent:   A,  monochtorinated adduct  of
pentachlorocyclopentadiene   with   cyclopentadiene;   B,   coelution   of
heptachlor and a-chlordene; C,  coelution of /3-chlordene and  7-chlordene;
D, a chlordane analog;  G, coelution of cis-nonachlor and "Compound K," a
chlordane isomer.   The  right  "shoulder" of peak  F is   caused by trans-
nonachlor.

            7.6.4.1     The GC pattern of a chlordane residue may differ
      considerably from that of the  technical  standard.  Depending on the
      sample substrate  and  its  history,  residues  of chlordane can consist
      of  almost   any  combination  of  constituents   from the  technical
      chlordane,   plant  and/or  animal  metabolites,  and  products  of
      degradation  caused by exposure  tc  environmental  factors  such  as
      water and sunlight.  Only limited  information is  available on which
      residue GC  patterns are likely to  occur  in  which  samples types,  and
      even this information may  not be applicable to a situation where the
      route of exposure  is  unusual.  For example,  fish exposed to a recent
      spill  of  technical chlordane will  contain a  residue  drastically
      different from a  fish  whose chlordane  residue was accumulated  by
      ingestion of smaller  fish  or of vegetation,   which  in turn  had
      accumulated  residues  because  chlordane  was  in  the  water  from
      agricultural runoff.

            7.6.4.2     Because of this inability to predict a chlordane
      residue GC  pattern, it is not  possible to prescribe a single method
      for the quantitation  of chlordane  residues.  The  analyst must judge
      whether or not the residue's GC pattern is sufficiently similar to
      that of a technical chlordane reference material to use the latter
      as a reference standard for quantitation.

            7.6.4.3     When  the  chlordane  residue  does  not  resemble
      technical chlordane,  but instead consists primarily of individual,
      identifiable peaks,  quantitate  each  peak   separately  against  the
      appropriate reference materials  and report  the  individual residues.
      (Reference  materials  are available  for  at   least  11  chlordane
      constituents, metabolites or degradation products which  may occur in
      the residue.)
                             8080A  -  9                         Revision 1
                                                          September 1994

-------
            7.6.4.4     When the  GC pattern of the residue resembles that
      of technical  chlordane, quantitate chlordane residues by comparing
      the total  area  of  the  chlordane chromatogram from peaks A through F
      (Figure 9a)  in the sample  versus the same  part  of  the  standard
      chromatogram.  Peak G may be obscured in a sample by the presence of
      other  pesticides.    If  G   is  not  obscured,   include  it  in  the
      measurement for both standard and sample.  If the heptachlor epoxide
      peak is relatively small, include  it as part of the total chlordane
      area  for   calculation  of   the  residue.     If  heptachlor  and/or
      heptachlor epoxide  are much out  of proportion as  in  Figure  6j,
      calculate  these separately  and subtract their areas from total area
      to give a  corrected chlordane  area.   (Note  that octachlor epoxide,
      a metabolite of chlordane,  can  easily be mistaken  for  heptachlor
      epoxide on a nonpolar GC column.)

            7.6.4.5     To  measure  the total   area  of  the  chlordane
      chromatogram,  proceed  as  in Sec.  7.6.2 on  toxaphene.   Inject  an
      amount  of  technical   chlordane   standard  which  will  produce  a
      chromatogram in which  peaks  E and  F are approximately the same size
      as  those  in the  sample chromatograms.    Construct  the  baseline
      beneath the standard from the beginning of peak  A to the end of peak
      F as shown in Figure 9a.  Use  the  distance  from the  trough between
      peaks E and F to the baseline in  the chromatogram of the standard to
      construct  the baseline in the chromatogram of the sample.  Figure 9b
      shows  how  the  presence of toxaphene  causes  the baseline  under
      chlordane  to take  an upward angle.  When the size  of peaks E  and F
      in standard and sample chromatograms are the  same, the distance from
      the  trough of the  peaks  to  the  baselines should  be  the  same.
      Measurement of  chlordane area  should be done by total  peak area if
      possible.

            NOTE: A  comparison  has  been  made  of the  total  peak  area
                  integration method  and  the addition  of  peak  heights
                  method for  several  samples containing chlordane.   The
                  peak   heights A,  B,  C,  D,  E,  and  F  were measured  in
                  millimeters from peak maximum of each to  the  baseline
                  constructed under the total chlordane area  and were then
                  added  together.   These  results obtained  by the  two
                  techniques are  too close to ignore this method of "peak
                  height addition" as a means of  calculating  chlordane.
                  The technique has inherent difficulties because not all
                  the peaks  are symmetrical and not all are present in the
                  same  ratio in standard and in sample.  This method does
                  offer  a means  of  calculating  results  if no  means  of
                  measuring total  area is  practical.

      7.6.5 Polychlorinated biphenyls  (PCBs): Quantitation of residues of
PCB involves problems similar to those encountered in the quantitation of
toxaphene, Strobane,  and chlordane.  In each case,  the chemical is made up
of numerous compounds.   So the chromatograms are multi-peak.  Also in each
case, the chromatogram of the residue may not match that of the standard.

            7.6.5.1     Mixtures  of PCBs of various chlorine  contents were
      sold  for  many  years  in the U.S.  by the  Monsanto  Co.  under  the

                            8080A -  10                        Revision 1
                                                          September 1994

-------
      tradename Aroclor (1200 series and 1016).  Though these Aroclors are
      no  longer  marketed,  the  PCBs  remain  in  the environment  and  are
      sometimes found as residues in foods, especially fish.

            7.6.5.2     PCB residues are quantitated by comparison to one
      or more of the Aroclor materials, depending on the chromatographic
      pattern of the residue.  A choice must be made as to which Aroclor
      or mixture of Aroclors will produce a chromatogram most similar to
      that of the residue.   This  may  also  involve a  judgment about what
      proportion of  the different  Aroclors to  combine  to  produce  the
      appropriate reference material.

            7.6.5.3     Quantitate PCB residues by comparing total area or
      height  of  residue peaks  to total  area of  height  of  peaks  from
      appropriate Aroclor(s) reference materials.  Measure total area or
      height  response  from common baseline  under all peaks.   Use only
      those   peaks   from  the   sample  that   can  be   attributed   to
      chlorobiphenyls.    These  peaks  must  also  be  present  in  the
      chromatogram of the reference materials.  Mixtures of Aroclors may
      be required to provide the best match of GC patterns of sample and
      reference.

      7.6.6 DDT:   DDT found  in  samples  often  consists of  both o,p'- and
p,p'-DDT.   Residues  of DDE  and  ODD  are also frequently  present.   Each
isomer of  DDT and  its metabolites should be  quantitated  using  the pure
standard of that compound and reported as such.

      7.6.7 Hexachlorocyclohexane   (BHC,  from the former  name,  benzene
hexachloride):   Technical  grade BHC is a  cream-colored  amorphous  solid
with a very  characteristic musty odor; it consists of a  mixture of six
chemically distinct isomers and one or more heptachloro-cyclohexanes and
octachloro-cyclohexanes.

            7.6,7.1     Commercial  BHC  preparations   may  show  a  wide
      variance in  the  percentage  of  individual  isomers  present.    The
      elimination rate of the isomers  fed to rats was 3 weeks for the a-,
      7-,  and 6-isomers and 14 weeks  for the 0-isomer.   Thus it may be
      possible to have any  combination of the various isomers in different
      food commodities.  BHC found in dairy products usually has a large
      percentage of /3-isomer.

            7.6.7.2     Individual isomers  (a, fi,  7, and s) were injected
      into   gas    chromatographs  equipped   with   flame   ionization,
      microcoulometric, and electron capture detectors.  Response for the
      four  isomers  is  very nearly the same whether flame  ionization or
      microcoulometric GLC is used.  The a-, 7-, and (5-isomers show equal
      electron affinity.   /3-BHC  shows  a  much weaker  electron  affinity
      compared to the other isomers.

            7.6.7.3     Quantitate  each   isomer   (a,   0,  7,   and   6)
      separately against a standard of the respective  pure isomer,  using
      a GC column which separates all  the isomers from one another.
                            8080A - 11                        Revision 1
                                                          September 1994

-------
8.0   QUALITY CONTROL

      8.1   Refer  to Chapter  One  for  specific  quality control  procedures.
Quality control to validate sample extraction is covered in Method 3500 and in
the extraction method utilized.  If extract cleanup was performed,  follow the QC
in Method 3600 and in the specific cleanup method.

      8.2   Quality control required to evaluate the GC system operation is found
in Method 8000.

            8.2.1 The quality  control  check sample  concentrate  (Method  8000)
      should  contain  each  single-component  parameter of   interest  at  the
      following  concentrations  in  acetone  or  other water  miscible  solvent:
      4,4'-ODD, 10 mg/L; 4,4'-DDT, 10 mg/L; endosulfan II,  10 mg/L; endosulfan
      sulfate,  10  mg/L;  endrin,   10 mg/L;  and  any  other  single-component
      pesticide, 2 mg/L.  If this method is only to be used to  analyze for PCBs,
      chlordane, or  toxaphene, the  QC  check sample concentrate should contain
      the most representative multi-component parameter at a concentration of 50
      mg/L in acetone.

            8.2.2 Table 3 indicates  the  QC acceptance criteria for this method.
      Table 4 gives method accuracy  and  precision  as  functions of concentration
      for the analytes of interest.   The contents  of  both Tables should be used
      to evaluate a laboratory's ability to perform and generate acceptable data
      by this method.

      8.3   Calculate surrogate standard  recovery on all  samples, blanks,  and
spikes.   Determine  if  the  recovery is  within limits  (limits  established by
performing QC procedures outlined in Method 8000).

            8.3.1 If recovery is not within limits, the following is required.

                  •     Check to be  sure  there are  no errors in calculations,
                        surrogate solutions and internal standards. Also, check
                        instrument performance.

                  •     Recalculate the data and/or reanalyze the extract if any
                        of the above checks reveal a problem.

                  •     Reextract and reanalyze the sample if none of the above
                        are  a  problem  or  flag   the  data  as  "estimated
                        concentration".

      8.4   GC/MS confirmation:  Any compounds confirmed by two columns may also
be confirmed by GC/MS if the  concentration is sufficient for detection by GC/MS
as determined by the laboratory generated detection limits.

            8.4.1 The GC/MS would normally require a minimum concentration of 10
      r\g/p,l  in the final extract, for each  single-component compound.

            8.4.2 The pesticide extract and associated blank  should be analyzed
      by GC/MS as per Sec. 7.0 of Method  8270.
                                  8080A - 12                        Revision 1
                                                                September 1994

-------
            8.4.3 The  confirmation may  be  from  the  GC/MS  analysis  of the
      base/neutral-acid extractables extracts (sample and blank).  However, if
      the  compounds  are  not  detected in  the base/neutral-acid  extract even
      though the concentration is high enough, a GC/MS analysis of  the pesticide
      extract should be performed.

            8.4.4 A reference standard of the compound must  also be analyzed by
      GC/MS.  The  concentration  of the reference  standard  must  be  at a  level
      that  would  demonstrate  the  ability  to  confirm  the  pesticides/PCBs
      identified by GC/ECD.


9.0   METHOD PERFORMANCE

      9.1   The method was tested by 20 laboratories  using organic-free reagent
water, drinking water,  surface water, and three industrial wastewaters  spiked at
six concentrations.  Concentrations used in the study ranged  from 0.5 to 30 /ug/L
for single-component pesticides  and from 8.5 to 400 ng/L  for multi-component
parameters.  Single operator precision, overall  precision,  and method accuracy
were  found  to  be directly related  to  the concentration of  the  parameter and
essentially independent of the sample matrix.  Linear equations to describe these
relationships for an electron capture detector are presented in Table 4.

      9.2   The accuracy and  precision obtained will be determined  by the  sample
matrix,   sample-preparation   technique,   optional    cleanup   techniques,  and
calibration procedures used.


10.0  REFERENCES

1.    U.S.  EPA,  "Development  and Application of Test Procedures  for Specific
      Organic  Toxic  Substances  in Wastewaters, Category  10:  Pesticides and
      PCBs," Report for EPA Contract 68-03-2605.

2.    U.S.  EPA,  "Interim Methods  for  the  Sampling  and Analysis  of Priority
      Pollutants  in  Sediments and   Fish Tissue,"  Environmental  Monitoring and
      Support Laboratory, Cincinnati, OH 45268,  October 1980.

3.    Pressley, T.A.,  and O.E.  Longbottom,  "The Determination of Organohalide
      Pesticides and PCBs in  Industrial and Municipal Wastewater: Method  617,"
      U.S.  EPA/EMSL, Cincinnati, OH, EPA-600/4-84-006, 1982.

4.    "Determination  of  Pesticides and  PCB's  in  Industrial  and  Municipal
      Wastewaters,   U.S.   Environmental   Protection  Agency,"   Environmental
      Monitoring and Support Laboratory, Cincinnati, OH 45268, EPA-600/4-82-023,
      June  1982.

5.    Goerlitz, D.F. and L.M.  Law,  Bulletin  for Environmental Contamination and
      Toxicology, 6, 9, 1971.

6.    Burke, J.A.,  "Gas Chromatography  for Pesticide Residue  Analysis; Some
      Practical  Aspects,"  Journal   of  the  Association of  Official  Analytical
      Chemists, 48, 1037, 1965.


                                  8080A - 13                        Revision 1
                                                                September 1994

-------
7.    Webb,  R.G.  and A.C.  McCall,  "Quantitative  PCB Standards  for Electron
      Capture Gas Chromatography," Journal  of Chromatographic Science, H, 366,
      1973.

8.    Millar, J.D.,  R.E.  Thomas  and H.J.  Schattenberg,  "EPA Method Study 18,
      Method 608: Organochlorine Pesticides and PCBs," U.S. EPA/EMSL, Research
      Triangle Park, NC, EPA-600/4-84-061, 1984.

9.    U.S. EPA 40 CFR Part 136, "Guidelines Establishing Test Procedures for the
      Analysis of Pollutants Under the Clean Water Act; Final  Rule and Interim
      Final Rule and Proposed Rule," October 26, 1984.

11.   U.S. Food  and  Drug  Administration,  Pesticide Analytical  Manual, Vol. 1,
      June 1979.

12.   Sawyer, L.D., JAOAC, 56, 1015-1023 (1973), 61 272-281 (1978), 61 282-291
      (1978).

13.   Stewart, J.  "EPA Verification Experiment for  Validation of the SOXTEC® PCB
      Extraction  Procedure";  Oak Ridge National  Laboratory,  Oak  Ridge,  TN,
      37831-6138; October 1988.
                                  8080A - 14                        Revision 1
                                                                September 1994

-------
                                   TABLE 1.
                  GAS CHROMATOGRAPHY OF PESTICIDES AND PCBs"
Analyte
Aldrin
a-BHC
jS-BHC
5-BHC
7-BHC (Lindane)
Chlordane (technical)
4,4'-DDD
4,4'-DDE
4,4'-DDT
Dieldrin
Endosulfan I
Endosulfan II
Endosulfan sulfate
Endrin
Endrin aldehyde
Heptachlor
Heptachlor epoxide
Methoxychlor
Toxaphene
PCB-1016
PCB-1221
PCB-1232
PCB-1242
PCB-1248
PCB-1254
PCB-1260
Retention
Col. 1
2.40
1.35
1.90
2.15
1.70
e
7.83
5.13
9.40
5.45
4.50
8.00
14.22
6.55
11.82
2.00
3.50
18.20
e
e
e
e
e
e
e
e
time (min)
Col. 2
4.10
1.82
1.97
2.20
2.13
e
9.08
7.15
11.75
7.23
6.20
8.28
10.70
8.10
9.30
3.35
5.00
26.60
e
e
e
e
e
e
e
e
Method
Detection
limit (M9/L)
0.004
0.003
0.006
0.009
0.004
0.014
0.011
0.004
0.012
0.002
0.014
0.004
0.066
0.006
0.023
0.003
0.083
0.176
0.24
nd
nd
nd
0.065
nd
nd
nd
aU.S.  EPA.   Method  617.   Organochlorine  Pesticides and  PCBs.
Monitoring and Support Laboratory, Cincinnati, Ohio 45268.

e  =  Multiple peak response.

nd = not determined.
 Environmental
                                  8080A - 15
    Revision 1
September 1994

-------
                                  TABLE 2.
DETERMINATION OF ESTIMATED QUANTITATION LIMITS (EQLs) FOR VARIOUS MATRICES8
        Matrix                                                Factor
        Ground water                                               10
        Low-concentration soil  by sonication with GPC cleanup     670
        High-concentration soil and sludges by sonication      10,000
        Non-water miscible waste                              100,000
           EQL = [Method detection limit (see Table 1)]  X  [Factor found in this
           table].   For  non-aqueous samples, the  factor is on  a  wet-weight
           basis.  Sample EQLs are  highly matrix-dependent.  The EQLs listed
           herein are provided for guidance  and  may not always  be achievable.
                                 8080A - 16                        Revision 1
                                                               September 1994

-------
                                   TABLE 3.
                            QC  ACCEPTANCE  CRITERIA3


Analyte
Aldrin
a-BHC
0-BHC
5-BHC
7-BHC
Chlordane
4,4'-DDD
4,4'-DDE
4,4'-DDT
Dieldrin
Endosulfan
Endosulfan
Endosulfan
Endrin
Heptachlor
Heptachlor
Toxaphene
PCB-1016
PCB-1221
PCB-1232
PCB-1242
PCB-1248
PCB-1254
PCB-1260
g =
x
P, Ps =
D
Test
cone.
(M9A)
2.0
2.0
2.0
2.0
2.0
50
10
2.0
10
2.0
I 2.0
II 10
Sulfate 10
10
2.0
epoxide 2.0
50
50
50
50
50
50
50
50
Standard deviation of
Average recovery for
Limit
for s
(M9/L)
0.42
0.48
0.64
0.72
0.46
10.0
2.8
0.55
3.6
0.76
0.49
6.1
2.7
3.7
0.40
0.41
12.7
10.0
24.4
17.9
12.2
15.9
13.8
10.4
four recovery
four recovery
Range
for x
(M9A)
1.08-2.24
0.98-2.44
0.78-2.60
1.01-2.37
0.86-2.32
27.6-54.3
4.8-12.6
1.08-2.60
4.6-13.7
1.15-2.49
1.14-2.82
2.2-17.1
3.8-13.2
5.1-12.6
0.86-2.00
1.13-2.63
27.8-55.6
30.5-51.5
22.1-75.2
14.0-98.5
24.8-69.6
29.0-70.2
22.2-57.9
18.7-54.9
measurements, in /
measurements, in ^
Range
P> PS
(%)
42-122
37-134
17-147
19-140
32-127
45-119
31-141
30-145
25-160
36-146
45-153
D-202
26-144
30-147
34-111
37-142
41-126
50-114
15-178
10-215
39-150
38-158
29-131
8-127

3/ *
Percent recovery measured.
Detected; result must
be greater than zero.
"Criteria  from 40 CFR Part 136 for Method 608.  These criteria are based directly
upon the method performance  data in Table  4.   Where necessary,  the limits for
recovery  have  been  broadened  to  assure applicability  of  the  limits  to
concentrations below those used to develop Table 4.
                                  8080A - 17
    Revision 1
September 1994

-------
                                   TABLE 4.
         METHOD ACCURACY AND PRECISION AS FUNCTIONS OF CONCENTRATION"
Analyte
Aldrin
a-BHC
0-BHC
5-BHC
7-BHC
Chlordane
4,4'-DDD
4,4'-DDE
4,4'-DDT
Dieldrin
Endosulfan I
Endosulfan II
Endosulfan Sulfate
Endrin
Heptachlor
Heptachlor epoxide
Toxaphene
PCB-1016
PCB-1221
PCB-1232
PCB-1242
PCB-1248
PCB-1254
PCB-1260
Accuracy, as
recovery, x'
(M9/L)
0.81C+0.04
0.84C+0.03
0.81C+0.07
0.81C+0.07
0.82C-0.05
0.82C-0.04
0.84C+0.30
0.85C+0.14
0.93C-0.13
0.90C+0.02
0.97C+0.04
0.93C+0.34
0.89C-0.37
0.89C-0.04
0.69C+0.04
0.89C+0.10
0.80C+1.74
0.81C+0.50
0.96C+0.65
0.91C+10.79
0.91C+10.79
0.91C+10.79
0.91C+10.79
0.91C+10.79
Single analyst
precision, s/
(M9/L)
0.16X-0.04
0.13X+0.04
0.22X+0.02
0.18X+0.09
0.12X+0.06
0.13X+0.13
O.ZOx-0.18
0.13X+0.06
0.17X+0.39
0.12X+O.I9
O.lOx+0.07
0.41X-0.65
0.13X+0.33
0.20X+0.25
0.06X+0.13
O.lSx-0.11
0.09X+3.20
0.13X+0.15
0.29X-0.76
0.21X-1.93
O.Zlx-1.93
0.21X-1.93
0.21X-1.93
fl.21x-l.93
Overall
precision,
S' (M9/L)
0.20X-0.01
0.23X-0.00
0.33X-0.95
0.25X+0.03
0.22x-t-0.04
O.lSx+0.18
0.27X-0.14
0.28X-0.09
0.31X-0.21
0.16X+0.16
0.18X+0.08
0.47X-0.20
0.24X+0.35
0.24X+0.25
0.16X+0.08
0.25X-0.08
0.20X+0.22
0.15X+0.45
0.35X-0.62
0.31X+3.50
0.31X+3.50
0.31X+3.50
0.31X+3.50
0.31X+3.50
X'
S'


C

x
Expected  recovery  for  one   or  more  measurements  of  a  sample
containing concentration C, in /Lig/L.

Expected single  analyst  standard deviation of measurements  at an
average concentration of x, in M9/L-

Expected interlaboratory standard deviation of measurements  at an
average concentration found of x, in
True value for the concentration, in /zg/L.

Average recovery  found  for measurements  of samples  containing  a
concentration of C, in M9A-
                                  8080A - 18
                                                        Revision 1
                                                    September 1994

-------
               Figure 1
      Gas Chromatogram of  Pesticides
Column:    1.5% SP-2250+
           1.95% SP-2401 on Supelcoport
Temperature:  200°C
Detector:  Electron Capture
              •        12
       MfTiNTION TIME (MINUTES)
              8080A - 19
    Revision 1
September 1994

-------
            Figure 2
  Gas Chromatogram of Chlordane
Column:    1.5% SP-2250+
           1.95% SP-2401  on Supelcoport
Temperature:  200°C
Detector:  Electron Capture
     4        I        12

   RETENTION TIME (MINUTES)
           8080A - 20
    Revision 1
September 1994

-------
         Figure  3
Gas  Chromatogram of Toxaphene
          Column:    1.5% SP-2250+
                     1.95% SP-2401  on Supelcoport
          Temperature:  200°C
          Detector:  Electron Capture
   10       H      II
 HfTtNTION TIMI (MINUTIS)
22
       8080A - 21
          Revision  1
       September 1994

-------
                 Figure 4
      Gas Chromatogram of Aroclor 1254
Column:    1.5% SP-2250+
           1.95% SP-2401  on Supelcoport
Temperature:  200°C
Detector:  Electron Capture
       6        10        14
         ftCTINTlON TIME (MINUTIS)
II
22
               8080A -  22
          Revision 1
      September 1994

-------
                      Figure 5
           Gas Chromatogram of Aroclor 1260
Column:    1.5% SP-2250+
           1.95% SP-2401  on Supelcoport
Temperature:  200°C
Detector:  Electron Capture
               10      14      II
             ftfTINTtON T1MI (MINUTIS)
22
2f
                      8080A - 23
             Revision 1
          September 1994

-------
                            Figure 6
        J..L
Fig.6--Baseline construction for some typical gas chromotagraphic peaks.
a: symmetrical separated flat baseline;  b and c:  overlapp flat  baseline;
d: separated (pen does not return to baseline between peaks); e: separated
sloping baseline; f:  separated (pen goes below baseline between peaks);
g: a- and 7-BHC  sloping baseline;  h: a-,ft-  and 7-BHC sloping baseline;
i: chlordane flat baseline;  j: heptachlor and heptachlor epoxide super-
imposed on chlordane;  k: chair-shaped peaks,  unsymmetrical peak;
1: p,p'-DDT superimposed on toxaphene.
                           8080A - 24
    Revision 1
September 1994

-------
                                Figure  7
 Fig.- 7a - - Baseline construction for multiple residues with standard
                               toxaphene.

Fig.- 7b -- Baseline construction for multiple residues with toxaphene,
                      DDE and o,p'-,  and p,p'-DDT
                               8080A  -  25
    Revision 1
September 1994

-------
                                  Figure 8
  Fig.- 8a -- Baseline construction for multiple residues: standard toxaphene.
Pig.- 8b -- Baseline construction for multiple residues:  rice bran with BHC,
                     toxaphene, DDT, and methoxychlor.
                               8080A - 26
    Revision 1
September 1994

-------
                               Figure 9
Pig.- 9a -- Baseline construction for multiple residues: standard chlordane.
Fig.- 9b -- Baseline construction for multiple residues: rice bran with
                     chlordane,  toxaphene,  and DDT.
                              8080A - 27
    Revision 1
September 1994

-------
                                METHOD  8080A
       ORGANOCHLORINE  PESTICIDES AND  POLYCHLORINATED BIPHENYLS
                           BY GAS  CHROMATOGRAPHY
       Start
     7.1.1 Choos«
appropriate extraction
      procedure.
   7.1.2 Exchange
  extraction solvent
     to hexane.
       7.2 Set
   chromatographic
      conditions.
      7.3 Refer to
   Method 8000 for
  proper calibration
      techniques.
 7.3.2 Prime or
deactivate the GC
 column prior to
 daily calibration.
  7.4 Perform
  GC analysis.
     7.4.8
     is peak
  detection and
   identification
   prevented?
     7.6.1 Do
  residues have
   two or more
   components?
   7.5.1 Cleanup
 using Method 3620
or 3660 if necessary.
    7.6 Calculate
   concentrations.
                                 8080A - 28
                                 Revision  1
                             September 1994

-------
00
o
00

-------
                                  METHOD  8081

             ORGANOCHLORINE  PESTICIDES  AND  PCBs  AS  AROCLORS  BY  GAS
                  CHROMATOGRAPHY: CAPILLARY COLUMN TECHNIQUE
1.0   SCOPE AND APPLICATION

      1.1   Method  8081  is used  to determine  the  concentrations  of  various
organochlorine pesticides and polychlorinated biphenyls (PCBs) as Aroclors, in
extracts from solid and liquid matrices.   Open-tubular, capillary columns were
employed with  electron capture detectors  (ECD) or  electrolytic  conductivity
detectors (ELCD).  When compared to the packed columns,  these fused-sil ica, open-
tubular  columns  offer  improved  resolution,   better  selectivity,  increased
sensitivity, and faster analysis.  The list below is annotated to show whether a
single- or  dual-column  analysis system was used to identify each target analyte.
            Compound Name
CAS Registry No.
Aldrin8-"
Aroclor-1016a'b
Aroclor-1221a'b
Aroclor-1232a'b
Aroclor-1242a'b
Aroclor-1248a-b
Aroclor-1254a'b
Aroclor-1260a'b
a-BHCa'b
/3-BHCa'b
-y-BHC (Lindane)a-b
5-BHCa-b
Chlorobenzilateb
a-Chlordaneb
T-Chlordanea-b
DBCPb
4,4'-DDDa'b
4,4'-DDEa'b
4,4'-DDTa'b
Diallate6
Dieldrina-b
Endosulfan Pb
Endosulfan IPb
Endosulfan sulfatea'b
Endrina'b
Endrin aldehydea'b
Endrin ketoneb
309-00-2
12674-11-2
1104-28-2
11141-16-5
53469-21-9
12672-29-6
11097-69-1
11096-82-5
319-84-6
319-85-7
58-89-9
319-86-8
510-15-6
5103-71-9
5103-74-2
96-12-8
72-54-8
72-55-9
50-29-3
2303-16-4
60-57-1
959-98-8
33213-65-9
1031-07-8
72-20-8
7421-93-4
53494-70-5
                                   8081  -  1
                          Revision  0
                      September  1994

-------
            Compound Name                 CAS Registry No.
             Heptachlora'b                     76-44-8
             Heptachlor epoxide"'b            1024-57-3
             Hexachlorobenzeneb               118-74-1
             Hexachlorocyclopentadieneb       77-47-4
             Isodrinb                         465-73-6
             Keponeb                          143-50-0
             Methoxychlora'b                   72-43-5
             Toxaphenea'b                     8001-35-2
                Single-column analysis
                Dual-column analysis
      1.2   The analyst must select columns, detectors and calibration procedures
most appropriate  for  the specific analytes  of  interest in a  study.   Matrix-
specific performance data must be established  and the stability  of the analytical
system and instrument  calibration must be established for each analytical matrix
(e.g.,  hexane solutions from sample extractions,  diluted oil samples, etc.).

      1.3   Although  performance  data  are  presented for  many of  the listed
chemicals,  it  is  unlikely that  all  of them  could  be determined  in  a single
analysis.   This limitation results  because  the chemical  and  chromatographic
behavior  of many  of  these  chemicals  can  result  in   co-elution.    Several
cleanup/fractionation schemes are  provided in  this  method  and in Method 3600.
Any chemical is a  potential method  interference when it is not a target analyte.

      1.4   Several multi-component mixtures  (i.e.,  Aroclors and Toxaphene) are
listed as target compounds.  When  samples  contain more than  one multi-component
analyte, a  higher  level  of  analyst expertise is  required to attain acceptable
levels of  qualitative  and  quantitative  analysis.   The  same is  true of multi-
component  analytes  that  have been  subjected to environmental  degradation or
degradation by treatment  technologies.  These result in "weathered" Aroclors (or
any other multi-component mixtures) that may have significant differences in peak
patterns than those of standards.   In  these cases, individual congener analyses
may be preferred over total mixture analyses.

      1.5   Compound  identification based on  single  column  analysis should be
confirmed  on  a second column,  or  should be  supported  by   at  least  one other
qualitative technique. This method describes analytical  conditions  for  a second
gas chromatographic column that can be used to confirm the measurements made with
the primary column.   GC/MS Method 8270 is also recommended as a confirmation
technique if sensitivity permits (Sec. 8).

      1.6   This method  describes  a  dual  column option.  The  option allows a
hardware configuration of two analytical  columns joined to a  single injection
port.    The option allows one injection  to  be used  for  dual  column analysis.

                                   8081 - 2                         Revision 0
                                                                September 1994

-------
Analysts are cautioned that the dual column option may not be appropriate when
the instrument is subject to mechanical  stress,  many  samples  are to be  run in a
short period, or when contaminated samples are analyzed.

      1.7   This method  is restricted to  use  by or  under  the  supervision of
analysts  experienced  in  the  use  of  a  gas chromatograph   (GC)  and  in  the
interpretation of gas chromatograms.   Each  analyst must demonstrate the ability
to generate acceptable results with this method.

      1.8   Extracts suitable for analysis by this method may also  be  analyzed
for  organophosphorus pesticides  (Method 8141).   Some  extracts  may  also  be
suitable for triazine  herbicide  analysis,  if low recoveries  (normally samples
taken for triazine analysis must be preserved) are not a problem.

      1.9   The following compounds may  also be determined using this method:
            Compound Name
CAS Registry No.
             Alachlor8-6
             Captafolb
             Captanb
             Chloroneb6
             Chloropropylateb
             Chlorothalonilb
             DCPAb
             Dichlone6
             Dicofolb
             Etridiazoleb
             Halowax-1000b
             Halowax-1001b
             Halowax-1013b
             Halowax-1014b
             Halowax-1051b
             Halowax-1099b
             Mirexb
             Nitrofenb
             PCNB"
             Perthaneb
             Propachlor6
             Strobaneb
             trans-Nonachlorb
             trans-Permethrinb
             Trifluralinb
 15972-60-8
  2425-06-1
   133-06-2
  2675-77-6
 99516-95-7
  1897-45-6
  1861-32-1
   117-80-6
   115-32-2
  2593-15-9
 58718-66-4
 58718-67-5
 12616-35-2
 12616-36-3
  2234-13-1
 39450-05-0
  2385-85-5
  1836-75-5
    82-68-8
    72-56-0
 1918-16-17
  8001-50-1
 39765-80-5
 51877-74-8
  1582-09-8
                Single-column analysis
                Dual-column analysis
                                   8081 - 3
                          Revision 0
                      September 1994

-------
2.0   SUMMARY OF METHOD

      2.1   A measured volume or weight of sample (approximately 1 L for 1 iquids,
2 g to 30  g  for solids) is extracted  using  the  appropriate sample extraction
technique.   Liquid samples are extracted at neutral pH with methylene chloride
using either  a  separatory  funnel  (Method 3510) or a  continuous liquid-liquid
extractor (Method 3520).  Solid  samples are extracted with hexane-acetone (1:1)
or methylene  chloride-acetone  (1:1)  using  either Soxhlet  extraction  (Method
3540), Automated Soxhlet (Method 3541), or Ultrasonic Extraction  (Method 3550).
A variety of cleanup steps may be applied to the extract, depending on (1) the
nature of  the coextracted matrix  interferences  and (2) the  target analytes.
After cleanup,  the extract  is  analyzed by  injecting a  1-^1-  sample  into  a gas
chromatograph with  a  narrow- or wide-bore  fused   silica capillary  column  and
electron capture  detector (GC/ECD)  or an  electrolytic  conductivity  detector
(GC/ELCD).


3.0   INTERFERENCES

      3.1   Refer to Methods 3500 (Sec. 3,  in particular),  3600, and 8000.

      3.2   Sources of  interference  in this method can  be  grouped  into  three
broad categories:  contaminated solvents, reagents or sample processing hardware;
contaminated GC  carrier  gas, parts,  column surfaces  or detector surfaces; and the
presence of  coeluting compounds in  the sample matrix  to  which the  ECD will
respond.  Interferences coextracted  from the  samples will  vary considerably from
waste to waste.   While general cleanup techniques are referenced  or provided as
part of this  method, unique samples may require additional cleanup approaches to
achieve desired degrees of discrimination and quantitation.

      3.3   Interferences  by   phthalate  esters   introduced   during   sample
preparation  can pose  a major  problem  in  pesticide  determinations.    These
materials  may be  removed  prior to  analysis using  Gel  Permeation Cleanup  -
pesticide  option  (Method  3640)  or as  Fraction  III of  the  silica  gel  cleanup
procedure  (Method 3630).  Common flexible  plastics contain  varying amounts of
phthalate esters which are easily extracted or leached from such materials during
laboratory operations.  Cross-contamination  of clean glassware routinely occurs
when plastics are handled during  extraction steps, especially when solvent-wetted
surfaces are handled.   Interferences from phthalate esters can best be minimized
by avoiding  contact with  any plastic  materials and checking  all solvents  and
reagents for phthalate contamination.   Exhaustive cleanup of solvents,  reagents
and  glassware  may  be  required  to  eliminate   background  phthalate  ester
contamination.

      3.4   Glassware must be scrupulously cleaned.  Clean all glassware as soon
as possible  after  use  by  rinsing  with the last solvent  used.   This should be
followed by  detergent washing with hot  water,  and rinses  with  tap water  and
organic-free reagent water.   Drain  the glassware and dry  in  an oven  at  130°C for
several hours or rinse with  methanol  and drain.  Store dry glassware in a clean
environment.

      3.5   The presence of  elemental  sulfur will result in  broad  peaks that
interfere with the detection of early-eluting organochlorine  pesticides.  Sulfur


                                   8081 - 4                         Revision 0
                                                                September 1994

-------
contamination should be expected with sediment samples.  Method 3660 is suggested
for removal  of  sulfur.   Since the recovery  of  Endrin  aldehyde (using  the TBA
procedure)  is drastically  reduced,  this compound must  be  determined  prior to
sulfur cleanup.

      3.6   Waxes, lipids, and other high molecular weight co-extractables can
be removed by Gel-Permeation Cleanup (Method 3640).

      3.7   It  may be  difficult to  quantitate Aroclor  patterns  and  single
component  pesticides  together.    Some  pesticides can  be removed  by  sulfuric
acid/permanganate cleanup (Method 3665}  and  silica fractionation (Method 3630).
Guidance on the identification of PCBs Is given in Sec. 7.

      3.8   The following target  analytes coelute using single column analysis:

            DB 608      Trifluralin/Diallate isomers
                        PCNP/Dichlone/Isodrin
                        DDD/Endosulfan II

            DB 1701     Captan/Chlorobenzilate
                        Captafol/Mirex
                        DDD/Endosulfan II
                        Methoxychlor/Endosulfan sulfate

            3.8.1 Other  halogenated  pesticides  or   industrial  chemicals  may
      interfere  with   the   analysis  of  pesticides.     Certain  co-eluting
      organophosphorus  pesticides   are  eliminated   by   the  Gel  Permeation
      Chromatography  cleanup  -   pesticide  option (Method  3640).    Co-eluting
      chlorophenols are  eliminated by Silica gel  (Method 3630), Florisil  (Method
      3620), or Alumina (Method  3610) cleanup.

      3.9   The following compounds coelute using the dual  column analysis.  Two
temperature programs are provided for the same pair  of columns as  option 1 and
option 2 for dual column analysis.  In general,  the  DB-5 column resolves fewer
compounds that the DB-1701:

            3.9.1 DB-5/DB-1701,  thin film,  slow ramp:  See Sec. 7  and  Table 6.

                  DB-5  trans-Permethrin/Heptachlor  epoxide
                        Endosulfan I/a-Chlordane
                        Perthane/Endrin
                        Endosulfan II/Chloropropylate/Chiorobenzi1 ate
                        4,4'-DDT/Endosulfan sulfate
                        Methoxychlor/Dicofol

            Perthane/Endrin and  Chiorobenzilate/Endosulfan  II/Chloropropylate
      will  also  co-elute  on  DB-5  after   moderate  deterioration  in  column
      performance.
                                   8081  - 5                         Revision 0
                                                                September 1994

-------
                  DB-1701     Chlorothalonil/8-BHC
                              £-BHC/DCPA/tra/7s-Permethrin
                              a-Chlordane/tra/?s-Nonachlor
                              Captan/Dieldrin
                              Chiorobenzi1 ate/Chioropropylate

            Chlorothalonil/G-BHC and or-Chlordane/trans-Nonachlor will co-elute
      on the DB-1701 column  after moderate deterioration in column performance.

            Nitrofen,  Dichlone,  Carbophenothion,  Dichloran  and Kepone  were
      removed from the composite mixture  because of  extensive peak tailing on
      both  columns.    Simazine  and Atrazine  give poor  responses  on  the  ECD
      detector.  Triazine compounds should  be  analyzed  using Method 8141 (NPD
      option).

            3.9.2 DB-5/DB-1701, thick film,  fast ramp:   See Sec.  7 and Table 7.

                  DB-5        Diallate/o-BHC
                              Perthane/Endosulfan II
                              Chiorobenzi1 ate/Chioropropylate
                              Endrin/Nitrofen
                              4,4'-DDT/Endosulfan sulfate
                              Methoxychlor/Dicolfol

                  DB-1701     a-Chlordane/ferans-Nonachlor {partially resolved)
                              4,4'-DDD/Endosulfan II  (partially resolved)


4.0   APPARATUS AND MATERIALS

      4.1   Gas  chromatograph:  an   analytical  system  complete   with  gas
chromatograph  suitable for  on-column  and  split-splitless  injection  and  all
required accessories  including  syringes,  analytical  columns,  gases, electron
capture detectors (ECD),  and recorder/integrator or data system.

      The  columns  listed  in this  section  were  used   to  develop  the  method
performance data.  Their specification is not intended  to prevent laboratories
from  using  columns  that  are  developed  after  promulgation  of the  method.
Laboratories may use other capillary columns  if they document method performance
data (e.g.  chromatographic resolution,  analyte  breakdown,  and MDLs) equal to or
better than those provided with the method.

            4.1.1 Single-column Analysis:

                  4.1.1.1     Narrow-bore columns:

                        4.1.1.1.1    Column 1 - 30 m x 0.25 or 0.32 mm internal
                  diameter (ID) fused  silica capillary column chemically bonded
                  with SE-54 (DB 5  or equivalent), 1  /urn film thickness.

                        4.1.1.1.2    Column 2 -  30 m  x 0.25 mm ID fused silica
                  capillary  column  chemically  bonded with  35  percent  phenyl
                                   8081 - 6                         Revision 0
                                                                September 1994

-------
      methylpolysiloxane  (DB  608,  SPB 608, or  equivalent),  25 pm
      coating thickness, 1 jitm film thickness.

            4.1.1.1.3   Narrow bore columns should be installed in
      split/splitless (Grob-type) injectors.

      4.1.1.2     Wide-bore columns

            4.1.1.2.1   Column 1 - 30 m  x 0.53  mm ID fused silica
      capillary column  chemically bonded  with  35  percent phenyl
      methylpolysiloxane (OB 608, SPB 608, RTx-35, or equivalent),
      0.5 jLtm or 0.83 urn film  thickness.

            4.1.1.2.2   Column 2 - 30 m  x  0.53  mm ID fused silica
      capillary column  chemically bonded  with  50  percent phenyl
      methylpolysiloxane  (DB  1701,  or  equivalent),  1.0  jum  film
      thickness.

            4.1.1.2.3   Column 3 - 30 m  x  0.53  mm ID fused silica
      capillary column chemically bonded  with  SE-54 (DB 5, SPB 5,
      RTx5, or equivalent), 1.5 ^m film thickness.

            4.1.1.2.4   Wide-bore columns should be installed in 1/4
      inch injectors, with deactivated liners designed specifically
      for use with these columns.

4.1.2 Dual Column Analysis:

      4.1.2.1     Column pair 1 :

            4.1.2.1.1   J&W Scientific press-fit Y-shaped glass 3-
      way union splitter (J&W Scientific, Catalog no. 705-0733) or
      Restek Y-shaped fused-silica connector  (Restek,  Catalog no.
      20405), or equivalent.

            4.1.2.1.2   30 m  x  0.53 m  ID DB-5  (J&W Scientific),
      1.5 /iiti film thickness,  or equivalent.

            4.1.2.1.3   30 m  x 0.53 mm ID DB-1701 (J&W Scientific),
      1.0 fj.m film thickness,  or equivalent.

      4.1.2.2     Column pair 2:

            4.1.2.2.1   Splitter 2 -  Supelco 8 in.  glass injection
      tee,   deactivated   (Supelco,   Catalog   no.    2-3665M),   or
      equivalent.

            4.1.2.2.2   30 m  x  0.53 m  ID OB-5  (J&W Scientific),
      0.83 /Ltm film thickness,  or equivalent.

            4.1.2.2.3   30 m  x 0.53 mm ID DB-1701 (J&W Scientific),
      1.0 y,m film thickness,  or equivalent.
                       8081 - 7                         Revision 0
                                                    September 1994

-------
            4.1.3 Column  rinsing  kit:  Bonded-phase  column  rinse  kit  (J&W
      Scientific, Catalog no.  430-3000 or  equivalent).

      4.2   Glassware (see Methods  3510,  3520,  3540,  3541, 3550,  3630,  3640,
3660,  and 3665 for specifications).

      4.3   Kuderna-Danish (K-D) apparatus. See extraction methods for specifics.


5.0   REAGENTS

      5.1   Reagent or  pesticide grade  chemicals  shall  be used in  all  tests.
Unless otherwise indicated, it is  intended that all reagents  shall  conform to
specifications of the Committee on  Analytical Reagents of the American Chemical
Society,  where such  specifications are  available.   Other grades may  be  used,
provided it is first  ascertained that the reagent is of sufficiently high purity
to permit its use without lessening the accuracy  of the  determination.

      NOTE: Store  the  standard   solutions  (stock,   composite,   calibration,
            internal, and surrogate) at 4°C in  Teflon-sealed  containers in the
            dark.  When a lot  of  standards is prepared,  it is  recommended that
            aliquots of that lot  be stored  in  individual  small  vials.  All  stock
            standard  solutions must  be replaced  after  one  year or  sooner  if
            routine  QC   (Sec.   8)  indicates  a  problem.   All  other  standard
            solutions must be  replaced after  six months or sooner if routine QC
            (Sec. 8) indicates a  problem.

      5.2   Solvents and reagents:   As appropriate for Method 3510,  3520,  3540,
3541,  3550,  3630,   3640,  3660,   or 3665:   n-hexane,  diethyl   ether,  methylene
chloride, acetone,  ethyl acetate, and isooctane (2,2,4-tnmethylpentane).  All
solvents should  be  pesticide  quality or  equivalent,  and each lot  of solvent
should be determined  to  be phthalate free.   Solvents must be exchanged to hexane
or isooctane prior to analysis.

            5.2.1 Organic-free reagent water:  All  references  to water in this
      method refer to organic-free  reagent water  as  defined  in Chapter One.

      5.3   Stock standard solutions  (1000 mg/L): Can  be prepared from pure
standard materials or can be  purchased as  certified  solutions.

            5.3.1 Prepare stock standard solutions by accurately weighing about
      0.0100 g of pure compound.   Dissolve the  compound  in isooctane or hexane
      and dilute to volume in  a 10-mL volumetric  flask.   If  compound purity is
      96 percent or  greater,  the  weight   can  be  used  without  correction  to
      calculate the concentration of the stock standard solution.   Commercially
      prepared stock standard  solutions  can be used at any concentration if they
      are certified by the manufacturer or by an  independent  source.

            5.3.2 6-BHC,  Dieldrin,  and  some  other   standards   may  not  be
      adequately soluble  in isooctane.  A small  amount of acetone  or toluene
      should be used to dissolve  these compounds  during  the  preparation of the
      stock standard solutions.
                                   8081 - 8                         Revision 0
                                                                September 1994

-------
      5.4   Composite stock  standard:  Can  be  prepared from  individual  stock
solutions.  For composite stock  standards  containing  less than 25 components,
take exactly 1 ml of each individual stock solution at 1000 mg/L, add solvent,
and mix the solutions in  a 25-mL volumetric flask.   For example, for a composite
containing 20  individual standards, the resulting concentration of  each component
in the mixture,  after the volume  is adjusted to 25 ml,  will be  1 mg/25 ml.  This
composite solution can be further diluted to obtain the desired concentrations.
For composite  stock standards containing more than  25 components,  use volumetric
flasks of the  appropriate volume  (e.g.,  50 ml, 100 ml).

      5.5   Calibration  standards should  be  prepared  at a  minimum of  five
concentrations by  dilution  of  the composite stock standard with isooctane or
hexane.    The concentrations  should  correspond  to   the expected  range  of
concentrations found in real  samples and  should bracket the linear range of the
detector.

            5.5.1 Although all single component analytes  can  be  resolved  on a
      new  35   percent  phenyl  methyl  silicone   column   (e.g.,  DB-608),   two
      calibration mixtures  should be prepared  for  the  single component analytes
      of this  method.

            5.5.2 This  procedure  is  established  to   (1) minimize  potential
      resolution and quantitation problems on confirmation columns or on older
      35  percent  phenyl  methyl  silicone  (e.g.  DB-608) columns  and  (2)  allow
      determination of Endrin and DDT breakdown for method QC (Sec.  8).

            5.5.3 Separate calibration standards  are  required for each  multi-
      component target analyte,  with the  exception of Aroclors 1016 and 1260,
      which can be  run as a mixture.

      5.6   Internal standard (optional):

            5.6.1 Pentachloronitrobenzene is suggested as an internal standard
      for the single column analysis, when it is not  considered to be a target
      analyte.   l-Bromo-2-nitrobenzene   is  a suggested  option.   Prepare  the
      standard to complement the  concentrations found  in  Sec.  5.5.

            5.6.2 Make a solution of 1000 mg/L of  l-bromo-2-nitrobenzene  for
      dual-column  analysis.   Dilute it  to  500  ng/>L  for spiking,  then use a
      spiking volume of  10 nl/ml  of extract.

      5.7   Surrogate  standards:   The  performance of the  method   should  be
monitored  using  surrogate  compounds.   Surrogate standards  are  added  to  all
samples, method blanks, matrix spikes,  and calibration standards.

            5.7.1 For the single column analysis,  use  decachlorobiphenyl as the
      primary surrogate.  However, if recovery is  low,  or late-eluting compounds
      interfere with decachlorobiphenyl,  then tetrachloro-m-xylene  should be
      evaluated  as  a  surrogate.   Proceed  with  corrective  action  when  both
      surrogates are out of limits for a sample (Sec.  8.2).  Method 3500,  Sec.
      5,  indicates  the proper procedure for preparing  these surrogates.
                                   8081 - 9                         Revision 0
                                                                September 1994

-------
            5.7.2 For the dual column analysis make a solution of 1000 mg/L of
      4-chloro-3-nitrobenzotrifluoride and dilute to 500 ng//iL.  Use a spiking
      volume of 100 /iL  for  all  aqueous  sample.   Store the spiking solutions
      at 4°C in Teflon-sealed containers  in  the dark.
6.0   SAMPLE COLLECTION, PRESERVATION AND HANDLING

      6.1   See Chapter 4, Organic Analytes, Sec. 4.

      6.2   Extracts must be stored under refrigeration in the dark and analyzed
within 40 days of extraction.


7.0   PROCEDURE

      7.1   Extraction:

            7.1.1 Refer to Chapter Two and  Method 3500 for guidance in choosing
      the  appropriate  extraction procedure.   In  general,  water  samples  are
      extracted at  a neutral  pH with methylene  chloride  as a  solvent using a
      separatory funnel  (Method  3510}  or  a continuous liquid-liquid extractor
      (Method 3520).  Extract solid  samples with  hexane-acetone  (1:1) using one
      of the Soxhlet extraction  (Method 3540 or 3541) or ultrasonic extraction
      (Method 3550) procedures.

            NOTE: Hexane/acetone  (1:1) may be  more  effective  as an extraction
                  solvent  for  organochlorine  pesticides and  PCBs  in  some
                  environmental   and  waste   matrices   than   is   methylene
                  chloride/acetone  (1:1).    Use  of hexane/acetone  generally
                  reduces the amount of co-extracted  interferences and improves
                  signal/noise.

            7.1.2 Spiked  samples  are used  to verify the  applicability  of  the
      chosen extraction  technique to each  new sample type.   Each  sample type
      must be  spiked with the compounds  of interest  to  determine  the percent
      recovery and the  limit of detection for that sample  (Sec.  5).  See Method
      8000 for guidance on demonstration  of initial  method proficiency as well
      as guidance on matrix spikes for routine sample analysis.

      7.2   Cleanup/Fractionation:

            7.2.1 Cleanup procedures may not be necessary for a relatively clean
      sample matrix, but most  extracts from environmental  and waste samples will
      require  additional  preparation before analysis.   The  specific cleanup
      procedure used will depend on  the nature  of the sample to  be analyzed and
      the data quality  objectives for  the  measurements.   General  guidance  for
      sample extract cleanup  is provided  in this section and in Method 3600.

                  7.2.1.1     If  a sample  is of biological origin,  or contains
            high molecular weight materials, the use of GPC  cleanup/pesticide
            option  (Method 3640)  is  recommended.    Frequently,   one of  the
                                   8081  -  10                         Revision 0
                                                                September 1994

-------
            adsorption chromatographic cleanups may also be required following
            the GPC cleanup.

                  7.2.1.2     If only PCBs are to be measured in a sample, the
            sulfuric  acid/permanganate  cleanup (Method 3665)  is recommended.
            Additional cleanup/fractionation by Alumina Cleanup  (Method 3610),
            Silica-Gel Cleanup (Method 3630), or Florisil Cleanup (Method 3620),
            may be necessary.

                  7.2.1.3     If both PCBs and  pesticides are to be measured in
            the sample, isolation of the PCB  fraction by Silica  Cleanup (Method
            3630) is recommended.

                  7.2.1.4     If only pesticides  are to be measured, cleanup by
            Method 3620 or Method 3630 is recommended.

                  7.2.1.5     Elemental   sulfur,  which  may  appear  in  certain
            sediments  and  industrial  wastes,  interferes   with the  electron
            capture gas chromatography of certain  pesticides.  Sulfur should be
            removed by the technique described in Method 3660,   Sulfur Cleanup.

      7.3   GC Conditions:     This method allows  the analyst to choose between
a single column  or  a  dual column configuration in the  injector port.   Either
wide- or narrow-bore  columns  may  be  used.   Identifications  based on retention
times from  a  single  column  must  be confirmed on  a  second  column  or  with an
alternative qualitative technique.

            7.3.1 Single Column Analysis:

                  7.3.1.1     This capillary GC/ECD method  allows  the analyst
            the option of using 0.25-0.32 mm  ID capillary columns (narrow-bore)
            or 0.53 mm ID capillary columns (wide-bore).   Performance data are
            provided   for  both  options.     Figures   1-6  provide   example
            chromatograms.

                  7.3.1.2     The use of narrow-bore columns  is recommended when
            the  analyst  requires  greater chromatographic  resolution.   Use of
            narrow-bore columns is suitable for relatively clean samples or for
            extracts that have  been  prepared with  one  or more  of the clean-up
            options referenced in the method.  Wide-bore columns (0.53 mm) are
            suitable for more complex environmental and waste matrices.

                  7.3.1.3     For the single column method  of  analysis,  using
            wide-bore capillary columns, Table 1  lists  average   retention times
            and method detection limits  (MDLs)  for  the target analytes in water
            and soil matrices.  For the  single  column method of  analysis,  using
            narrow-bore capillary columns, Table 2 lists average  retention times
            and method detection limits  (MDLs)  for  the target analytes in water
            and soil matrices.  The MDLs  for the components of a  specific sample
            may  differ  from those listed  in Tables  1  and   2 because  they are
            dependent  upon  the  nature of interferences in  the  sample matrix.
            Table 3  lists the Estimated Quantitation Limits (EQLs)  for  other
            matrices.  Table 4 lists the GC operating conditions  for the single
            column method of analysis.

                                   8081  - 11                         Revision 0
                                                                September 1994

-------
      7.3.2 Dual Column Analysis:

            7.3.2.1     The dual-column/dual-detector  approach involves
      the use of two 30 m x 0.53 mm ID fused-silica open-tubular columns
      of different  polarities,  thus  different  selectivities towards the
      target compounds.  The columns are connected to an injection tee and
      ECD detectors.  Retention times for the organochlorine analytes on
      dual  columns  are in Table  5.   The  GC  operating conditions for the
      compounds in  Table  5  are in Table 6.  Multicomponent mixtures of
      Toxaphene and Strobane were  analyzed  separately  (Figures 7  and 8)
      using the GC operating conditions found in Table 7.  Seven Aroclor
      mixtures and six  Halowax mixtures were  analyzed under the conditions
      outlined in Table 7 (Figures 9 through 21).  Figure 22 is a sample
      chromatogram  for  a  mixture  of  organochlorine  pesticides.    The
      retention times  of  the  individual components  detected  in  these
      mixtures are given in Tables 8 and 9.

                  7.3.2.1.1   Operating  conditions  for  a   more  heavily
            loaded DB-5/DB-1701 pair are given  in Table 7.   This  column
            pair   was   used   for  the   detection   of  multicomponent
            organochlorine compounds.

                  7.3.2.1.2   Operating  conditions   for a  DB-5/DB-1701
            column pair with  thinner films,  a different  type of splitter,
            and a  slower temperature  programming  rate  are  provided  in
            Table  6.    These  conditions gave  better  peak shapes  for
            compounds  such as  Nitrofen and  Dicofol.   Table  5  lists the
            retention  times  for the compounds  detected on  this  column
            pair.

7.4   Calibration:

      7.4.1 Prepare calibration standards using  the procedures  in Sec. 5.
Refer to Method 8000 (Sec. 7)  for proper calibration techniques for both
initial   calibration and  calibration  verification.    The procedure  for
either  internal or external  calibration may be used,   however,  in  most
cases external  standard  calibration  is used with Method 8081.  This  is
because  of  the sensitivity  of  the  electron capture  detector and  the
probability of  the internal  standard  being affected  by interferences.
Because  several  of the  pesticides may  co-elute  on any single  column,
analysts should use two calibration mixtures  (see Sec. 3.8).  The specific
mixture should be selected to minimize  the  problem of peak  overlap.

      NOTE:  Because of the sensitivity  of the electron  capture detector,
            the injection port  and column should always be  cleaned  prior
            to performing the initial calibration.

            7.4.1.1     Method   8081  has  many  multi-component  target
      analytes.    For   this  reason,   the   target  analytes  chosen  for
      calibration  should  be  limited to  those  specified in  the  project
      plan.    For  instance,  some  sites may  require  analysis for  the
      organochlorine pesticides only or the  PCBs only.   Toxaphene  and/or
      technical Chlordane  may  not  be  specified at  certain sites.   In
      addition, where PCBs are specified in  the  project  plan, a mixture of

                            8081  -  12                         Revision 0
                                                          September 1994

-------
      Aroclors 1016 and 1260 will suffice for the initial calibration of
      all  Aroclors,  since  they  include  all  congeners  present  in  the
      different regulated Aroclors.   A mid-point calibration standard of
      all  Aroclors  (for Aroclor pattern recognition)  must  be  included with
      the initial calibration so that  the  analyst  is familiar with each
      Aroclor pattern and retention times on each column.

            7.4.1.2     For calibration verification  (each  12  hr shift)
      all  target analytes required  in  the  project  plan  must be injected
      with the  following exception  for  the Aroclors.   For  sites  that
      require PCB analysis,  include only the Aroclors that are expected to
      be found at the site.  If PCBs are required, but it  is unknown which
      Aroclors may be present, the  mid-concentration Aroclors 1016/1260
      mixture only, may be  injected.   However,  if  specific  Aroclors are
      found  at the  site during the initial  screening,  it  is required that
      the samples containing Aroclors be reinjected with the proper mid-
      concentration Aroclor standards.

      7.4.2  Because  of  the  low  concentration   of   pesticide  standards
injected on  a GC/ECD,  column adsorption may be a problem when the GC has
not been  used for  a day  or more.    Therefore,  the  GC column  should be
primed or deactivated by  injecting  a PCB  or  pesticide  standard mixture
approximately  20 times  more  concentrated than  the  mid-concentration
standard.   Inject  this  standard  mixture prior  to  beginning the initial
calibration  or calibration verification.

      CAUTION:    Several  analytes,  including  Aldrin, may be observed in
                  the  injection  just  following  this   system  priming.
                 Always  run  an  acceptable blank  prior  to  running  any
                  standards or samples.

      7.4.3  Retention time windows:

            7.4.3.1     Before establishing the  retention  time windows,
      make sure the gas chromatographic system  is within optimum operating
      conditions.  The width of the  retention time window should be based
      upon actual retention times of standards  measured over  the course of
      72 hours.  See Method 8000  for details.

            7.4.3.2     Retention time windows shall be defined  as plus or
      minus  three times  the standard deviation of the absolute retention
      times  for  each standard.   However,  the experience of the analyst
      should  weigh heavily  in  the   interpretation of the chromatograms.
      For multicomponent standards  (i.e.,  PCBs), the analyst should use
      the retention  time window but should  primarily  rely on  pattern
      recognition.   Sec.  7.5.4 provides guidance on the  establishment of
      absolute retention  time windows.

            7.4.3.3     Certain analytes, particularly Kepone, are subject
      to changes in retention  times.  Dry Kepone  standards  prepared in
      hexane  or  isooctane  can  produce  gaussian  peaks.   However,  Kepone
      extracted from samples or standards exposed to  water or methanol may
      produce peaks with broad tails that  elute later than  the standard
      (0-1 minute).  This  shift is presumably the result of the formation

                            8081 -  13                         Revision 0
                                                          September 1994

-------
      of a  hemi-acetal  from the  ketone  functionality.   Method  8270 is
      recommended for Kepone.

7.5   Gas chromatographic analysis:

      7.5.1 Set up the GC system using the conditions described in Tables
4, 6,  or 7.  An initial oven temperature at or below 140-150°C is required
to resolve  the  four BHC  isomers.   A final  temperature  of 240-270°C is
required  to  elute   decachlorobiphenyl.     Use   of  injector  pressure
programming will  improve the chromatography of late eluting peaks.

      7.5.2 Verify calibration  each 12 hour shift  by injecting calibration
verification standards prior to conducting any analyses.   See Sec. 7.4.1.2
for special guidance on calibration verification  of PCBs.  A calibration
standard must also be  injected at  intervals  of not less  than  once every
twenty  samples  (after every 10 samples  is recommended  to  minimize the
number of samples requiring  re-injection when  QC  limits are exceeded) and
at the  end  of the analysis  sequence.   The calibration  factor  for each
analyte to be quantitated must not exceed  a  +15  percent  difference when
compared to  the  initial calibration  curve.   When  this  criterion  is
exceeded, inspect the  gas chromatographic  system to determine the cause
and perform whatever maintenance is necessary  before verifying calibration
and proceeding with  sample   analysis.   If routine  maintenance  does not
return the instrument performance  to  meet the QC  requirements  {Sec. 8.2}
based on the last initial calibration, then a  new  initial calibration must
be performed.

            7.5.2.1     Analysts should use high and low concentrations of
      mixtures of single-component analytes and multi-component analytes
      for calibration verification.

      7.5.3 Sample injection may continue for as  long as  the calibration
verification standards and  standards  interspersed  with the samples meet
instrument QC  requirements.   It  is  recommended that  standards be analyzed
after every 10 (required  after every 20 samples),  and at the end of a set.
The sequence  ends when  the set  of  samples   has  been  injected  or when
qualitative and/or quantitative QC criteria are exceeded.

            7.5.3.1     Each sample  analysis  must be bracketed  with an
      acceptable initial  calibration,  calibration  verification standard(s)
      (each 12 hr shift), or calibration  standards interspersed within the
      samples.   All  samples that  were injected  after the  standard that
      last met the QC criteria  must be reinjected.

            7.5.3.2     Although analysis of  a  single  mid-concentration
      standard (standard mixture or multi-component analyte) will satisfy
      the  minimum  requirements,  analysts are  urged  to use  different
      calibration    verification    standards    during    organochlorine
      pesticide/PCB analyses.  Also,  multi-level  standards  (mixtures or
      multi-component  analytes)  are  highly  recommended  to ensure that
      detector  response  remains   stable  for  all  analytes   over  the
      calibration range.
                             8081  -  14                         Revision 0
                                                          September 1994

-------
      7.5.4 Establish absolute retention  time  windows for each analyte.
Use the absolute retention time for each analyte from standards analyzed
during  that  12 hour  shift  as the  midpoint of  the  window.    The daily
retention  time window equals  the midpoint  +  three  times  the standard
deviations.

            7.5.4.1     Tentative identification  of an analyte occurs when
      a peak from a sample extract falls within the daily retention time
      window.

            7.5.4.2     Validation   of   gas   chromatographic   system
      qualitative  performance:  Use  the  calibration  standards analyzed
      during the sequence  to evaluate retention time stability.   If any of
      the standards fall  outside their daily retention time windows,  the
      system is out  of  control.  Determine the cause  of  the  problem  and
      correct  it.

      7.5.5 Record the volume injected to  the  nearest  0.05 yuL and  the
resulting  peak size  in area  units.    Using  either  the  internal  or  the
external calibration procedure (Method 8000), determine the identity  and
the quantity  of each  component  peak  in  the sample  chromatogram which
corresponds to the compounds used for calibration purposes.

            7.5.5.1     If the responses exceed the calibration range of
      the  system,  dilute  the   extract   and   reanalyze.   Peak  height
      measurements  are  recommended   over  peak  area  integration  when
      overlapping peaks cause errors in area integration.

            7.5.5.2     If partially overlapping or  coeluting  peaks  are
      found, change  columns  or try  GC/MS  quantitation,  see Sec.  8  and
      Method 8270.

            7.5.5.3     If the peak  response is  less  than  2.5  times  the
      baseline noise  level, the validity of the  quantitative result may be
      questionable.  The  analyst  should  consult with the  source  of  the
      sample to determine whether further concentration of the  sample is
      warranted.

      7.5.6 Identification of mixtures {i.e.  PCBs and Toxaphene) is based
on  the  characteristic  "fingerprint"  retention  time  and  shape of  the
indicator  peak(s);  and  quantitation  is  based  on  the  area  under  the
characteristic  peaks  as   compared  to the  area  under the  corresponding
calibration peak(s) of the same retention time  and  shape  generated using
either internal or external  calibratibn procedures.

      7.5.7 Quantitation   of  the  target compounds  is  based  on:   1)    a
reproducible response of the ECD or ELCD within the calibration range;  and
2}  a  direct  proportionality between  the magnitude  of  response  of  the
detector to peaks  in  the  sample extract  and the calibration standards.
Proper quantitation requires  the appropriate  selection of a baseline from
which  the area  or  height of the characteristic peak(s) can be determined.
                            8081  -  15                         Revision 0
                                                          September 1994

-------
      7.5.8 If compound identification or quantitation is precluded due to
interference  (e.g.,  broad,  rounded  peaks  or ill-defined  baselines  are
present) cleanup of the extract  or  replacement of the capillary column or
detector  is  warranted.    Rerun the sample  on another  instrument  to
determine if  the problem results from  analytical hardware  or the sample
matrix.  Refer to Method 3600  for the procedures to be followed in sample
cleanup.

7.6   Quantitation of Multiple Component Analytes:

      7.6.1 Multi-component analytes present  problems   in  measurement.
Suggestions are offered in the following sections for handing Toxaphene,
Chlordane, PCBs, DDT, and BHC.

      7.6.2 Toxaphene:  Toxaphene is manufactured by the chlorination of
camphenes, whereas Strobane results from the chlorination of a mixture of
camphenes and pinenes.  Quantitative  calculation of Toxaphene or Strobane
is difficult,  but reasonable  accuracy  can be  obtained.   To  calculate
Toxaphene on  GC/ECD:   (a)    adjust  the sample size  so that  the  major
Toxaphene peaks are  10-70% of full-scale deflection (FSD);  (b)   inject a
Toxaphene standard that  is  estimated to be within  ±10  ng of the sample;
(c)  quantitate  using  the  five major  peaks  or  the total  area of  the
Toxaphene pattern.

            7.6.2.1     To measure  total area,  construct the baseline of
      standard  Toxaphene  between  its  extremities;  and construct  the
      baseline under the sample, using the distances of the peak troughs
      to  baseline  on the standard  as a guide.   This procedure  is  made
      difficult by the  fact that the relative  heights  and  widths of the
      peaks in the sample will  probably  not be identical to the standard.

            7,6.2.2     A   series   of   Toxaphene  residues   have   been
      calculated using the  total peak area for comparison to the standard
      and also using the area  of the  last four peaks only, in both sample
      and standard.  The agreement between  the results obtained  by the two
      methods justifies the  use of the  latter  method  for  calculating
      Toxaphene  in  a  sample  where  the early  eluting  portion of  the
      Toxaphene  chromatogram  shows  interferences from  other  substances
      such as DDT.

      7.6.3 Chlordane  is  a  technical  mixture  of  at   least  11  major
components and 30 or more minor components.   Trans- and C7's-Chlordane (a
and 7, respectively), are the two major components of technical Chlordane.
However,  the  exact  percentage of  each in  the technical  material  is  not
completely defined, and is not consistent  from  batch to batch.

            7.6.3.1     The GC pattern of  a Chlordane residue may differ
      considerably from that of the technical standard.  Depending on the
      sample substrate and  its history,  residues of Chlordane can consist
      of  almost  any  combination of:   constituents  from  the  technical
      Chlordane,  plant  and/or  animal   metabolites,   and   products  of
      degradation  caused  by exposure to  environmental  factors  such  as
      water and sunlight.
                             8081  -  16                         Revision 0
                                                          September 1994

-------
            7.6.3.2     Whenever possible, when a Chlordane residue does
      not resemble technical  Chlordane,  the analyst  should quantitate the
      peaks of a-Chlordane, 7-Chlordane, and Heptachlor separately against
      the  appropriate reference  materials,  and  report  the  individual
      residues.

            7.6.3.3     When the GC pattern of the residue resembles that
      of  technical  Chlordane,   the  analyst   may   quantitate  Chlordane
      residues by comparing the total area of the Chlordane chromatogram
      using the  five  major  peaks or the total  area.   If the Heptachlor
      epoxide  peak is  relatively  small,  include it  as part  of the total
      Chlordane area for calculation of the residue.   If Heptachlor and/or
      Heptachlor  epoxide  are much out  of proportion,  calculate  these
      separately and  subtract their areas  from the  total  area to give a
      corrected  Chlordane  area.     (Note   that  octachloro  epoxide,  a
      metabolite  of  Chlordane,   can easily be mistaken  for  Heptachlor
      epoxide  on a nonpolar GC column.)

            7.6.3.4     To  measure  the  total  area   of  the  Chlordane
      chromatogram, inject an amount of technical Chlordane standard which
      will  produce   a  chromatogram   in   which   the   major  peaks  are
      approximately the same size as those in  the sample chromatograms,

      7.6.4 Polychlorinated biphenyls (PCBs):  Quantitation of residues of
PCBs involves  problems similar to those encountered in  the quantitation of
Toxaphene, Strobane,  and Chlordane.   In each case,  the  material is made up
of numerous compounds which generate multi-peak chromatograms.   Also,  in
each case,  the chromatogram of  the residue may  not  match  that of  the
standard.

             7.6.4.1    Mixtures of PCBs of various chlorine contents were
      sold for many years  in  the  U.S. by the Monsanto Co. under the trade
      name Aroclor (1200 series and  1016).   Although  these Aroclors are no
      longer  marketed, the  PCBs   remain  in   the   environment  and  are
      sometimes found as residues in foods,  especially  fish.  The Aroclors
      most commonly found  in the environment are 1242, 1254,  and 1260.

            7.6.4.2     PCB  residues   are  generally  quantitated   by
      comparison to the most similar Aroclor standard.  A choice must  be
      made as  to which Aroclor is most similar  to that  of the residue and
      whether  that standard  is  truly representative of the PCBs  in  the
      sample.

            7.6.4.3     PCB  Quantitation option  #1- Quantitate the  PCB
      residues by  comparing  the  total  area of  the  chlorinated  biphenyl
      peaks to the total  area  of  peaks   from  the  appropriate  Aroclor
      reference material.  Measure  the total area or height response from
      the common baseline  under  all  the peaks.   Use only those peaks from
      the sample that can  be attributed to chlorobiphenyls.   These  peaks
      must also be present in the chromatogram of the reference materials.
      Option #1 should not be used  if there are interference  peaks  within
      the Aroclor pattern, especially if they  overlap  PCB  congeners.
                            8081 - 17                         Revision  0
                                                          September 1994

-------
                  7.6.4.4     PCB Quantitation  option  #2- Quantitate  the  PCB
            residues by comparing the responses of 3 to  5 major  peaks in each
            appropriate  Aroclor  standard  with  the  peaks  obtained   from  the
            chlorinated biphenyls in  the  sample  extract.  The amount of Aroclor
            is calculated using an individual response factor  for  each of the
            major peaks.  The results of the 3  to 5 determinations are averaged.
            Major peaks are defined as those peaks  in the Aroclor standards that
            are at least 25% of the height of the  largest Aroclor peak.  Late-
            eluting  Aroclor  peaks   are   generally  the  most  stable  in  the
            environment.

                  7.6.4.5     When samples  appear  to  contain  weathered  PCBs,
            treated PCBs, or mixtures of  Aroclors, the use of Aroclor standards
            is not appropriate.  Several  diagnostic peaks useful for identifying
            non-Aroclor PCBs are  given  in Table 10.   Analysts  should examine
            chromatograms containing  these peaks carefully, as these samples may
            contain PCBs.   PCB concentrations may be estimated  from specific
            congeners by adding the concentration  of the  congener peaks listed
            in Table 11.  The congeners are analyzed as single components.  This
            approach will provide reasonable accuracy  for Aroclors  1016,  1232,
            1242 and 1248 but will underestimate the concentrations of Aroclors
            1254,  1260  and  1221.   It  is  highly  recommended  that  heavily
            weathered,  treated, or mixed Aroclors  be  analyzed using  GC/MS  if
            concentration permits.

            7.6.5 Hexachlorocyclohexane   (BHC,  from  the  former name,  benzene
      hexachloride):   Technical  grade  BHC is a cream-colored  amorphous  solid
      with a very  characteristic  musty  odor;  it consists of a mixture of  six
      chemically distinct isomers and one  or  more heptachlorocyclohexanes  and
      octachlorocyclohexanes.    Commercial  BHC  preparations may  show a  wide
      variance in  the  percentage  of  individual  isomers present.    Quantitate
      each  isomer (a,  /3,  7,  and  6)  separately  against  a  standard of  the
      respective pure isomer.

            7.6.6 DDT:   Technical DDT consists primarily  of a mixture of 4,4'-
      DDT  (approximately 75%)  and  2,4'-DDT  (approximately   25%).    As  DDT
      weathers, 4,4'-DDE, 2,4'-DDE,  4,4'-DDD, and 2,4'-DDD  are  formed.  Since
      the 4,4'-isomers  of  DDT, DDE,  and  ODD  predominate in the  environment,
      these  are   the   isomers  normally   regulated  by  US  EPA  and  should  be
      quantitated against standards of the respective  pure isomer.

      7.7   Suggested chromatography maintenance: Corrective measures may require
any one or more of the following remedial actions.

            7.7.1 Splitter connections:   For  dual  columns which  are  connected
      using a  press-fit Y-shaped  glass  splitter or  a Y-shaped  fused-silica
      connector (J&W Scientific or Restek), clean and  deactivate  the splitter
      port insert or replace with a cleaned and deactivated splitter.  Break off
      the first few inches (up to one foot) of  the injection port  side of  the
      column.    Remove  the  columns  and   solvent  backflush  according to  the
      manufacturer's instructions. If these procedures  fail to  eliminate  the
      degradation problem,  it  may  be  necessary to deactivate the metal injector
      body and/or replace the  columns.


                                  8081 -  18                         Revision 0
                                                                September 1994

-------
                  7.7.1.1     GC  injector ports  can  be of  critical  concern,
            especially in the analysis  of DDT  and Endrin.   Injectors that are
            contaminated,  chemically   active,   or   too  hot  can  cause  the
            degradation ("breakdown") of the analytes.  Endrin and  DDT breakdown
            to Endrin aldehyde,  Endrin ketone, ODD, or DDE.  When such breakdown
            is observed, clean  and  deactivate  the injector port,  break off at
            least 0.5  M of  the column  and  remount  it.   Check  the  injector
            temperature  and  lower  it to  205°C,  if  required.  Endrin  and DDT
            breakdown is less of a problem when  ambient  on-column  injectors are
            used.

            7.7.2 Metal  injector body:    Turn  off  the oven  and  remove the
      analytical  columns when the oven has cooled.  Remove the glass injection
      port insert (instruments with on-column injection).  Lower the injection
      port temperature  to  room  temperature.   Inspect  the injection  port and
      remove any noticeable foreign material.

                  7.7.2.1     Place a beaker beneath  the injector port inside
            the oven.  Using  a wash  bottle, serially rinse the entire inside of
            the injector port with acetone and then  toluene;  catch the rinsate
            in the beaker.

                  7.7.2.2     Prepare a  solution of a deactivating  agent (Sylon-
            CT or equivalent) following  manufacturer's  directions.   After all
            metal surfaces  inside the injector body have  been thoroughly coated
            with   the  deactivation   solution,  rinse  the  injector  body  with
            toluene, methanol,  acetone,  then hexane.   Reassemble  the  injector
            and replace the columns.

            7.7.3 Column rinsing:   The  column  should be rinsed  with  several
      column  volumes of  an  appropriate  solvent.   Both  polar   and  nonpolar
      solvents are recommended.   Depending on the  nature  of the sample residues
      expected,  the first  rinse might  be water,  followed  by   methanol  and
      acetone; methylene chloride is  a good final rinse and  in  some cases may be
      the  only solvent  required.    The  column  should  then  be  filled  with
      methylene  chloride and  allowed  to stand  flooded  overnight to  allow
      materials within the stationary phase to  migrate  into  the  solvent.   The
      column is then flushed with fresh methylene chloride,  drained, and dried
      at room temperature with a stream of ultrapure nitrogen.


8.0   QUALITY CONTROL

      8.1   Refer to Chapter One for  specific  quality control  (QC)  procedures
including matrix  spikes, duplicates  and blanks.  Quality control  to  validate
sample  extraction  is  covered   in Method 3500  and  in  the  extraction  method
utilized.  If an extract cleanup was performed,  follow  the QC in Method 3600 and
in the specific cleanup method.

      8.2   Quality control requirements for the GC system, including cal ibration
and corrective actions,  are found  in  Method 8000.    The following steps are
recommended as additional method QC.
                                  8081  -  19                         Revision 0
                                                                September 1994

-------
            8.2.1 The  QC  Reference  Sample  concentrate  (Method 3500}  should
      contain the organochlorine pesticides at 10 mg/L for water  samples.   If
      this  method  is  to  be  used  for  analysis  of  Aroclors,  Chlordane,  or
      Toxaphene  only,  the   QC  Reference  Sample  should   contain  the  most
      representative multi-component mixture at a concentration of 50 mg/L in
      acetone.   The frequency of analysis of the QC reference sample analysis is
      equivalent to  a minimum of 1 per 20 samples  or  1  per batch if  less than 20
      samples.   If the recovery of any compound  found in the QC reference sample
      is less  than  80 percent  or  greater than  120 percent  of the  certified
      value, the laboratory  performance  is judged  to be out of control, and the
      problem must  be corrected.  A new set of calibration  standards should be
      prepared and  analyzed.

            8.2.2 Calculate  surrogate standard recovery on all samples, blanks,
      and  spikes.     Determine  if   the   recovery   is  within  limits  (limits
      established by performing QC procedures outlined in Method 8000).

            If recovery is not within limits, the following are required:

                  8.2.2.1      Confirm that there are no errors in calculations,
            surrogate solutions and  internal  standards.  Also, check instrument
            performance.

                  8.2.2.2      Examine chromatograms for  interfering  peaks  and
            for integrated areas.

                  8.2.2.3      Recalculate the data and/or reanalyze the extract
            if any of the  above checks reveal a problem.

                  8.2.2.4      Reextract and reanalyze  the sample if none of the
            above are a problem or flag  the data  as  "estimated  concentration."

            8.2.3 Include  a  calibration  standard after each group of 20 samples
      (it is recommended that  a calibration  standard be included after every 10
      samples  to minimize the  number of  repeat  injections)  in  the  analysis
      sequence as a calibration check.  The response  factors for the calibration
      should be within 15  percent  of  the  initial calibration.    When  this
      continuing calibration is out  of  this  acceptance window,  the laboratory
      should stop analyses and take corrective  action.

            8.2.4 Whenever  quantitation  is  accomplished  using   an  internal
      standard,  internal  standards  must  be evaluated  for  acceptance.    The
      measured area  of the  internal  standard must  be no more  than 50 percent
      different from the  average area calculated  during  calibration.   When  the
      internal  standard peak area is  outside the  limit,  all  samples  that  fall
      outside the QC criteria must be reanalyzed.

      8.3   DDT and  Endrin are easily degraded in  the injection port.  Breakdown
occurs when the injection  port liner is  contaminated high boiling  residue  from
sample  injection  or  when the injector  contains  metal   fittings.   Check  for
degradation problems  by injecting a standard containing only 4,4'-DDT and Endrin.
Presence of 4,4'-DDE, 4,4'-ODD, Endrin ketone or Endrin indicates breakdown.  If
degradation of either DDT  or Endrin exceeds 15%,  take corrective action before
proceeding with calibration.

                                  8081  -  20                         Revision 0
                                                                September  1994

-------
            8.3.1 Calculate percent breakdown as follows:

            % breakdown    Total DDT degradation peak area  (DDE + ODD)
            for 4,4'-DDT = 	  x 100
                                    peak areas  (DDT + DDE + ODD)

                              Total endrin degradation peak area
            % breakdown       (Endrin aldehyde  + Endrin ketone)
            for Endrin   =     	       x 100
                              peak areas (Endrin + aldehyde + ketone)

            8.3.2 The  breakdown  of DDT  and  Endrin should  be  measured before
      samples are analyzed and at the beginning  of each 12 hour shift.  Injector
      maintenance and  recalibration  should  be  completed if the  breakdown is
      greater than 15% for either compound (Sec. 8.2.3).

      8.4   GC/MS confirmation  may be  used  for single  column  analysis.   In
addition, any compounds  confirmed  by two columns  should  also  be  confirmed by
GC/MS if the concentration is sufficient for detection by GC/MS.

            8.4.1 Full-scan GC/MS will  normally  require a minimum concentration
      near  10 ng//j,L in the final  extract  for  each single-component compound.
      Ion trap  or  selected   ion monitoring  will  normally  require  a  minimum
      concentration near 1 ng//xl_.

            8.4.2 The  GC/MS  must  be  calibrated  for  the  specific  target
      pesticides when  it is used for quantitative analysis.

            8.4.3 GC/MS may  not be used  for single  column  confirmation  when
      concentrations are below 1 ng/|xL.

            8.4.4 GC/MS confirmation  should  be accomplished by  analyzing  the
      same extract used for GC/ECD analysis and the associated blank.

            8.4.5 Use of the base/neutral-acid  extract and associated blank may
      be used if the surrogates and internal  standards do not interfere and it
      is demonstrated that the analyte  is stable during acid/base partitioning.
      However,  if the  compounds   are  not  detected  in   the  base/neutral-acid
      extract even though the concentrations are high enough, a GC/MS analysis
      of the pesticide extract should be performed.

            8.4.6 A QC reference sample of the compound must  also be analyzed by
      GC/MS.  The concentration of the  QC  reference  standard must demonstrate
      the ability to confirm the pesticides/Aroclors identified by GC/ECD.

      8.5   Whenever silica gel  (Method  3630) or Florisil (Method 3620) cleanup
is  used, the  analyst must  demonstrate  that   the   fractionation  scheme  is
reproducible.   Batch  to  batch variation in the composition  of  the  silica  gel
material or overloading  the  column  may  cause  a  change  in the  distribution
patterns of  the  organochlorine pesticides and PCBs.   When  compounds are found in
two fractions, add the concentrations  in  the  fractions, and corrections for any
additional  dilution.
                                   8081  -  21                         Revision 0
                                                                September 1994

-------
9.0   METHOD PERFORMANCE

      9.1   The MDL is defined  in Chapter One.  The MDL concentrations listed in
Tables 1 and  2  were obtained using organic-free  reagent water  and sandy loam
soil.

      9.2   The chromatographic separations in this method have been tested in
a single laboratory by  using clean  hexane  and liquid and solid waste extracts
that  were  spiked with  the  test compounds  at three concentrations.   Single-
operator precision,  overall  precision, and  method accuracy were  found  to be
related to the concentration of the compound and the type of matrix.

      9.3   This method has been applied in a variety of commercial laboratories
for environmental  and waste matrices.   Performance data were  obtained  for a
limited number of target analytes spiked into sewage sludge and dichloroethene
still  bottoms at high concentration levels.  These data are provided in Tables
12 and 13.

      9.4   The accuracy and  precision  obtainable with this method depend on the
sample matrix,  sample preparation  technique,  optional  cleanup techniques, and
calibration procedures  used.

      9.5   Single  laboratory  accuracy data were  obtained  for organochlorine
pesticides  in a clay soil.    The  spiking  concentration was  500 M9A9-   The
spiking solution was mixed  into the soil and -then immediately transferred to the
extraction device and immersed  in the extraction solvent.  The  spiked sample was
then  extracted by Method 3541 (Automated Soxhlet).  The data represent a single
determination.   Analysis was  by capillary  column gas chromatography/electron
capture detector following  Method 8081  for the organochlorine pesticides.  These
data  are listed  in Table 14 and were taken from Reference 14.

      9.6   Single  laboratory  recovery data were  obtained for PCBs  in clay and
soil.    Oak  Ridge  National   Laboratory  spiked  Aroclors  1254  and 1260  at
concentrations  of  5  and  50  ppm  into  portions of  clay  and  soil  samples and
extracted  these  spiked samples using  the  procedure outlined  in Method 3541.
Multiple extractions using two different extractors were performed.  The extracts
were  analyzed by Method 8081.   The data are listed in Table 15 and were taken
from  Reference  15.

      9.7   Multi-laboratory accuracy and precision data  were  obtained for PCBs
in  soil.   Eight  laboratories spiked Aroclors  1254  and 1260  into three portions
of  10 g of Fuller's  Earth on three non-consecutive  days, followed by immediate
extraction using Method 3541.  Six of the laboratories  spiked  each Aroclor at 5
and 50 mg/kg  and two  laboratories spiked each Aroclor at  50  and  500  mg/kg.  All
extracts were analyzed by  Oak  Ridge National Laboratory, Oak Ridge, TN, using
Method 8081.  These data are listed in Table 16 and were taken from Reference 13.
                                   8081 - 22                         Revision  0
                                                                 September  1994

-------
10.0  REFERENCES

1.    Lopez-Avila, V.;  Baldin,  E.;  Benedicto, J; Milanes,  J.;  Beckert.  W. F.
      Application of Open-Tubular Columns to SW-846 GC Methods";  final report to
      the U.S.  Environmental  Protection Agency  on Contract  68-03-3511;  Mid-
      Pacific Environmental Laboratory, Mountain View, CA, 1990.

2.    Development and Application of Test Procedures for Specific Organic Toxic
      Substances in Wastewaters.  Category  10 -  Pesticides and  PCB Report for
      the U.S. Environmental Protection Agency on Contract 68-03-2606.

3.    Goerlitz, D.F.;  Law,  L.M.  "Removal of Elemental  Sulfur Interferences from
      Sediment Extracts for Pesticide Analysis"; Bull.  Environ. Contam. Toxicol.
      1971,  6, 9.

4.    Ahnoff, M.; Josefsson, B.   "Cleanup Procedures for PCB Analysis on River
      Water Extracts"; Bull. Environ. Contam. Toxicol. 1975, 13, 159.

5.    Jensen, S.; Renberg,  L.; Reutergardth, L.  "Residue Analysis of Sediment
      and Sewage Sludge  for  Organochlorines in  the  Presence of  Elemental
      Sulfur"; Anal. Chem.  1977, 49, 316-318.

6.    Wise,  R.H.; Bishop, D.F.; Williams,  R.T.; Austern, B.M.   "Gel Permeation
      Chromatography  in the  GC/MS  Analysis of  Organics  in   Sludges";  U.S.
      Environmental  Research Laboratory.  Cincinnati,  OH  45268.

7.    Pionke, H.B.; Chesters,  G.; Armstrong,  D.E.   "Extraction  of Chlorinated
      Hydrocarbon Insecticides from Soil";  Agron. J.  1968, 60,  289.

8.    Burke, J.A.; Mills, P.A.; Bostwick, D.C.  "Experiments with Evaporation of
      Solutions of Chlorinated Pesticides"; J. Assoc.  Off.  Anal.  Chem. 1966, 49,
      999.

9.    Glazer, J.A.,  et al.   "Trace Analyses  for Wastewaters"; Environ. Sci. and
      Technol. 1981, 15, 1426.

10.   Marsden, P.J.,  "Performance Data for SW-846 Methods 8270, 8081, and 8141,"
      EMSL-LV, EPA/600/4-90/015.

11.   Marsden, P.O., "Analysis of PCBs", EMSL-LV, EPA/600/8-90/004

12.   Erickson,  M.  Analytical  Chemistry of  PCBs, Butterworth  Publishers,  Ann
      Arbor Science Book (1986).

13.   Stewart, J. "EPA Verification Experiment for Validation of the SOXTEC* PCB
      Extraction  Procedure";  Oak Ridge National Laboratory,  Oak  Ridge,  TN,
      37831-6138; October 1988.

14.   Lopez-Avila, V. (Beckert, W.,  Project Officer),  "Development of a Soxtec
      Extraction  Procedure for  Extracting  Organic Compounds  from  Soils  and
      Sediments", EPA  600/X-91/140,  US EPA,  Environmental  Monitoring  Systems
      Laboratory-Las Vegas, October 1991.
                                   8081  -  23                         Revision 0
                                                                September 1994

-------
15.    Stewart,  J.H.; Bayne, C.K.; Holmes,  R.L.;  Rogers,  W.F.;  and Maskarinec,
      M.P., "Evaluation of a Rapid Quantitative  Organic  Extraction System for
      Determining the Concentration of PCB in Soils", Proceedings of the USEPA
      Symposium  on  Waste Testing  and Quality  Assurance,  Oak  Ridge  National
      Laboratory, Oak Ridge,  TN 37831-6131; July 11-15, 1988.
                                   8081  -  24                         Revision 0
                                                                September 1994

-------
                           TABLE 1

  GAS CHROMATOGRAPHIC RETENTION TIMES AND  METHOD  DETECTION
LIMITS FOR THE ORGANOCHLORINE PESTICIDES AND PCBs AS AROCLORS
              USING WIDE-BORE CAPILLARY  COLUMNS
              SINGLE COLUMN METHOD OF ANALYSIS

Compound
Aldrin
ff-BHC
6-BHC
£-BHC
7-BHC (Lindane)
a-Chlordane
7-Chlordane
4,4'-DDD
4,4'-DDE
4,4'-DDT
Dieldrin
Endosulfan I
Endosulfan II
Endosulfan Sulfate
Endrin
Endrin aldehyde
Heptachlor
Heptachlor epoxide
Methoxychlor
Toxaphene
Aroclor-1016
Aroclor-1221
Aroclor-1232
Aroclor-1242
Aroclor-1248
Aroclor-1254
Aroclor-1260
Water = Organic-
Retention
DB 608b
11.84
8.14
9.86
11.20
9.52
15.24
14.63
18.43
16.34
19.48
16.41
15.25
18.45
20.21
17.80
19.72
10.66
13.97
22.80
MR
MR
MR
MR
MR
MR
MR
MR
free reagent
Time (min)
DB 1701b
12.50
9.46
13.58
14.39
10.84
16.48
16.20
19.56
16.76
20.10
17.32
15.96
19.72
22.36
18.06
21.18
11.56
15.03
22.34
MR
MR
MR
MR
MR
MR
MR
MR
water.
MDLa Water
(M9/L)
0.034
0.035
0.023
0.024
0.025
0.008
0.037
0.050
0.058
0.081
0.044
0.030
0.040
0.035
0.039
0.050
0.040
0.032
0.086
NA
0.054
NA
NA
NA
NA
NA
0.90

MDL° Soil
(M9/kg)
2.2
1.9
3.3
1.1
2.0

1.5
4.2
2.5
3.6
NA
2.1
2.4
3.6
3.6
1.6
2.0
2.1
5.7
NA
57.0
NA
NA
NA
NA
NA
70.0

Soil = Sandy loam soil .
MR = Multiple
NA = Data not
peak responses.
available.

MDL is the method detection limit. MDL

was determi

ned from
                                                               the
 analysis  of  seven  replicate  aliquots of  each  matrix processed
 through  the  entire  analytical   method  (extraction,   silica  gel
 cleanup, and GC/ECD analysis).  MDL = t(n-l, 0.99) x SD,  where  t(n-
 1, 0.99) is the Student's t value appropriate  for a 99%  confidence
 interval and  a  standard deviation with  n-1 degrees  of freedom, and
 SD is  the  standard deviation of  the seven replicate measurements.
 See Table  4 for GC operating conditions.
                       8081 - 25
    Revision 0
September 1994

-------
                                TABLE  2

        GAS CHROMATOGRAPHIC RETENTION TIMES AND METHOD DETECTION
      LIMITS  FOR  THE  ORGANOCHLORINE  PESTICIDES  AND  PCBs  AS  AROCLORS
                   USING  NARROW-BORE CAPILLARY  COLUMNS
                    SINGLE COLUMN METHOD OF ANALYSIS
Compound
Retention Time (min)
  DB 608b    DB 5b
MDLa Water MDLa Soil
 (M9/L)
Aldrin
o-BHC
B-BHC

-------
                          TABLE  2
                         (Continued)

MDL is  the  method detection  limit.   MDL was  determined  from the
analysis  of seven  replicate  aliquots  of  each matrix  processed
through  the entire  analytical method  (extraction,  cleanup,  and
GC/ECD analysis).  MDL = t(n-l, 0.99) x SD, where  t(n-l,  0.99) is
the Student's t value appropriate for a 99% confidence interval and
a standard  deviation with n-1 degrees  of freedom, and SD  is the
standard deviation of the seven replicate measurements.

30 m x 0.25  mm ID DB-608 1  pm film  thickness,  see Table  4  for GC
operating conditions.

30 m x  0.25 mm  ID DB-5  1 jitm  film  thickness,  see Table  4  for GC
operating conditions.
                      8081  -  27                         Revision 0
                                                    September 1994

-------
                                  TABLE 3

DETERMINATION OF ESTIMATED QUANTITATION LIMITS (EQLs) FOR VARIOUS MATRICES8



    Matrix                                                    Factor


    Ground water                                                   10
    Low-concentration soil by sonication with GPC cleanup         670
    High-concentration  soil and sludges  by  sonication          10,000
    Non-water miscible  waste                                  100,000
    EQL =  [Method detection limit for water  (see Table  1 or Table 2) wide-
    bore or narrow-bore options] x [Factor found in this table].  For
    nonaqueous samples, the factor is on a wet-weight basis.  Sample EQLs
    are highly matrix-dependent.  The EQLs to be determined herein are
    provided for guidance and may not always be achievable.
                                 8081  -  28                        Revision 0
                                                              September 1994

-------
                                   TABLE 4
             GC  OPERATING CONDITIONS FOR ORGANOCHLORINE COMPOUNDS
                            SINGLE  COLUMN ANALYSIS
      Narrow-bore  columns:

Narrow-bore Column 1 - 30 m x 0.25 or 0.32 mm  internal diameter  (ID)  fused
silica capillary column chemically bonded with SE-54  (DB-5 or  equivalent),  1
jum film thickness.
      Carrier gas  (He)
      Injector temperature
      Detector temperature
      Initial temperature
      Temperature  program

      Final temperature
16 psi
225°C
300°C
100°C,  hold 2 minutes
100°C  to 160°C  at  15°C/min, followed
by 160°C to 270°C  at  5°C/min
270°C
Narrow-bore Column 2 - 30 m x 0.25 mm ID fused silica capillary column
chemically bonded with 35 percent phenyl methylpolysiloxane  (DB-608,  SPB-608,
or equivalent), 25 jum coating thickness,  1  /xm film thickness
      Carrier gas  (N2)
      Injector temperature
      Detector temperature
      Initial temperature
      Temperature  program
      Final temperature
20 psi
225°C
300°C
160°C,
160°C
290°C,
 hold 2 minutes
to 290°C  at 5°C/min
 hold 1 min
      Wide-bore columns:

Wide-bore Column 1 - 30 m x 0.53 mm ID fused silica capillary column
chemically bonded with 35 percent phenyl methylpolysiloxane  (DB-608, SPB-608,
RTx-35,  or equivalent), 0.5 urn or 0.83 fj.m  film  thickness.

Wide-bore Column 2 - 30 m x 0.53 mm ID fused silica capillary column
chemically bonded with 50 percent phenyl methylpolysiloxane  (DB-1701, or
equivalent), 1.0 /urn  film thickness.
      Carrier gas  (He)
      Makeup gas
      argon/methane  (P-5  or  P-10)  or  N2
      Injector temperature
      Detector temperature
      Initial temperature
      Temperature  program
      Final temperature
5-7 mL/minute

30 mL/min
250°C
290°C
150°C,  hold  0.5  minute
150°C  to  270°C at  5°C/min
270°C,  hold  10 min
                                                                    (continued)
                                   8081  -  29
                          Revision 0
                      September 1994

-------
                             TABLE 4   (Continued)
             GC  OPERATING CONDITIONS  FOR ORGANOCHLORINE COMPOUNDS
                            SINGLE COLUMN ANALYSIS
Wide-bore Columns  (continued)

Wide-bore Column 3 - 30 m x 0.53 mm  ID  fused  silica  capillary  column
chemically bonded with SE-54  (DB-5,  SPB-5,  RTx-5,  or equivalent),  1.5 JLUTI film
thickness.
      Carrier  gas  (He)
      Makeup gas
      argon/methane  (P-5  or  P-10)  or  N2
      Injector temperature
      Detector temperature
      Initial  temperature
      Temperature  program
       Final  temperature
6 mL/minute

30 mL/min
205°C
290°C
140°C,  hold 2 min
140°C to 240°C  at 10°C/nriri»
hold 5 minutes at 240°C,
240°C to 265°C  at 5°C/min
265°C,  hold 18 min
                                   8081 - 30
                           Revision 0
                       September 1994

-------
                    TABLE 5
RETENTION TIMES OF THE ORGANOCHLORINE PESTICIDES"
         DUAL COLUMN METHOD  OF ANALYSIS
Compound
DBCP
Hexachl orocycl open tad i ene
Etridiazole
Chloroneb
Hexachl orobenzene
Diallate
Propachlor
Trifluralin
a-BHC
PCNB
y-BHC
Heptachlor
Aldrin
Alachlor
Chlorothalonil
Alachlor
0-BHC
Isodrin
DC PA
S-BHC
Heptachlor epoxide
Endosulfan-I
7-Chlordane
a-Chlordane
trans-Nonachlor
4,4'-DDE
Dleldrin
Captan
Perthane
Endrin
Chloropropylate
Chlorobenzllate
Nitrofen
4,4'-DDD
Endosulfan II
4,4'-DDT
Endrin aldehyde
Mi rex
Endosulfan sulfate
CAS No.
96-12-8
77-47-4
2593-15-9
2675-77-6
118-74-1
2303-16-4
1918-16-17
1582-09-8
319-84-6
82-68-8
58-89-9
76-44-8
309-00-2
15972-60-8
1897-45-6
15972-60-8
319-85-7
465-73-6
1861-32-1
319-86-8
1024-57-3
959-98-8
5103-74-2
5103-71-9
39765-80-5
72-55-9
60-57-1
133-06-2
72-56-0
72-20-8
99516-95-7
510-15-6
1836-75-5
72-54-8
33213-65-9
50-29-3
7421-93-4
2385-85-5
1031-07-8
DB-5
RT(min)
2.14
4.49
6.38
7.46
12.79
12.35
9.96
11.87
12.35
14.47
14.14
18.34
20.37
18.58
15.81
18.58
13.80
22.08
21.38
15.49
22.83
25.00
24.29
25.25
25.58
26.80
26.60
23.29
28.45
27.86
28.92
28.92
27.86
29.32
28.45
31.62
29.63
37.15
31.62
DB-1701
RT(min)
2.84
4.88
8.42
10.60
14.58
15.07
15.43
16.26
17.42
18.20
20.00
21.16
22.78
24.18
24.42
24.18
25.04
25.29
26.11
26.37
27.31
28.88
29.32
29.82
30.01
30.40
31.20
31.47
32.18
32.44
34.14
34.42
34.42
35.32
35.51
36.30
38.08
38.79
40.05
                                                      continued
                   8081 - 31
    Revision 0
September 1994

-------



Compound
Methoxychlor
Captafol
Endrin ketone
trans-Permethrin
Kepone
Dicofol
Dichlone
a, or -Dibromo-m-xylene
2-Bromobiphenyl
TABLE 5
(Continued)

CAS No.
72-43-5
2425-06-1
53494-70-5
51877-74-8
143-50-0
115-32-2
117-80-6




DB-5
RT(min)
35.33
32.65
33.79
41.50
31.10
35.33
15.17
9.17
8.54


DB-1701
RT(min)
40.31
41.42
42.26
45.81
b
b
b
11.51
12.49
3The GC operating conditions were as follows:  30-m x 0.53-mm  ID DB-5
 (0.83-jum film thickness) and 30-m x 0.53-mm  ID DB-1701  (1.0-jum film
 thickness) connected to an 8-in injection tee (Supelco  Inc.).  Temperature
 program: 140°C  (2-min hold)  to  270°C  (1-min  hold)  at   2.8°C/min; injector
 temperature 250°C;  detector  temperature 320°C; helium  carrier gas  6 mL/min;
 nitrogen makeup gas 20 mL/min.
bNot detected at 2 ng per injection.
                                   8081 - 32
    Revision 0
September 1994

-------
Column 1
                                    TABLE  6
              GC  OPERATING  CONDITIONS  FOR  ORGANOCHLORINE PESTICIDES
                      FOR DUAL COLUMN METHOD OF ANALYSIS
                          LOW TEMPERATURE, THIN FILM
            Type:  DB-1701 (J&W) or equivalent
            Dimensions:  30 m x 0.53 mm ID
            Film Thickness (urn):  1.0
Column 2;
            Type:  DB-5 (J&W) or equivalent
            Dimensions:  30 m x 0.53 mm ID
            Film Thickness (y^m):  0.83
Carrier gas flowrate (mL/min):  6 (Helium)
Makeup gas flowrate (mL/min):  20 (Nitrogen)
Temperature program:  140°C (2 min hold)  to 270°C  (1  min  hold)  at  2.8°C/min
Injector temperature:  250°C
Detector temperature:  320°C
Injection volume:  2 pi
Solvent:  Hexane
Type of injector:  Flash vaporization
Detector type:  Dual ECD
Range:  10
Attenuation:  64 (DB-1701)/32 (DB-5)
Type of splitter:  Supelco 8  in injection tee
                                   8081  -  33                        Revision 0
                                                                September 1994

-------
Column 1:
                                   TABLE 7
             GC OPERATING CONDITIONS FOR ORGANOCHLORINE PESTICIDES
                    FOR THE DUAL COLUMN METHOD OF ANALYSIS
                         HIGH TEMPERATURE, THICK FILM
            Type:  DB-1701 (J&W) or equivalent
            Dimensions:  30 m x 0.53 mm ID
            Film Thickness:  1.0 n\n
Column 2:
            Type:  DB-5 (J&W) or equivalent
            Dimensions:  30 m x 0.53 mm ID
            Film Thickness:  1.5 jum

Carrier gas flowrate (mL/min):  6 (Helium)

Makeup gas flowrate (mL/min):  20 (Nitrogen)

Temperature program:   150°C (0.5 min hold) to 190°C  (2  min  hold)  at  12°C/min
                       then to 275°C  (10 min hold)  at 4°C/imn.

Injector temperature:   250°C

Detector temperature:   320°C

Injection volume:  2 fj.1

Solvent:  Hexane

Type of injector:  Flash vaporization

Detector type:  Dual ECD

Range:  10

Attenuation:  64 (DB-1701J/64 (DB-5)

Type of splitter:  J&W Scientific press-fit  Y-shaped inlet splitter
                                   8081 - 34                         Revision 0
                                                                 September  1994

-------
              TABLE 8   SUMMARY OF RETENTION TIMES  (MIN) OF AROCLORS
                                ON THE OB-5 COLUMN8
                              DUAL SYSTEM OF ANALYSIS
Peak
No.6
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41

42
43
44
45
46
47
48

49
50
51
52
53
54
55
Aroclor
1016


8.41
8.77
8.98
9.71
10.49
10.58
10.90
11.23
11.88
11.99
12.27
12.66
12.98
13.18
13.61
13.80
13.96
14.48
14.63
14.99
15.35
16.01

16.27































Aroclor Aroclor
1221 1232
5.85 5.85
7.63 7.64
8.43 8.43
8.77 8.78
8.99 9.00

10.50 10.50
10.59 10.59
10.91
11.24 11.24
11.90
12.00
12.29 12.29
12.68 12.69
12.99 13.00
13.19
13.63
13.82
13.97
14.50
14.64
15.02
15.36

16.14
16.29

17.04
17.22
17.46



18.41
18.58

18.83
19.33


20.03





21.18










Aroclor
1242

7.57
8.37
8.73
8.94
9.66
10.44
10.53
10.86
11.18
11.84
11.95
12.24
12.64
12.95
13.14
13.58
13.77
13.93
14.46
14.60
14.98
15.32
15.96
16,08
16.26


17.19
17.43

17.92
18.16
18.37
18.56

18.80
19.30


19.97


20.46

20.85
21.14



22.08






Aroclor
1248




8.95

10.45

10.85
11.18
11.85

12.24
12.64
12.95
13.15
13.58
13.77
13.93
14.45
14.60
14.97
15.31

16.08
16.24

16.99
17.19
17.43
17.69
17,91
18.14
18.36
18.55

18.78
19.29


19.92


20.45

20.83
21.12
21,36


22.05






Aroclor
1254
















13.59
13.78
13.90
14.46

14.98
15.32

16.10
16.25
16.53
16.96
17.19
17.44
17.69
17.91
18.14
18.36
18.55

18.78
19.29
19.48
19.81
19.92

20.28

20.57
20.83
20.98
21.38
21.78

22.04
22.38
22.74
22.96
23.23

23.75
Aroclor
1260
















13.59








16.26

16.97
17.21




18.37

18.68
18.79
19.29
19.48
19.80


20.28

20.57
20.83

21.38
21.78

22.03
22.37
22.73
22.95
23.23
23.42
23.73
Pesticide eluting at same
retention time









Chlorothaloni I (11.18)















Captan (16.21)

gamma-Chlordane (16.95)





4,4'-DDE (18.38)
Dieldrin (18.59)





Chloropropylate (19.91)
Endosulfan II (19.91)




Kepone (20.99)

4,4'-ODT (21.75)
Endosulfan sulfate (21.75)


Captafol (22.71)



Endrin ketone (23.73)
'The GC operating conditions are given in Table 7.
                                                                            (continued)
                                     8081  - 35
    Revision 0
September  1994

-------
                                     TABLE 8   CONTINUED
Peak
No.
56
57
58

59
60
61
62
63
64
65
66
67
68
69
Aroclor Aroclor Aroclor Aroclor Aroclor Aroclor
1016 1221 1232 1242 1248 1254
23.99

24.27


24.61
24.93

26.22






Aroclor Pesticide eluting at same
1260 retention time
23.97
24.16
Methoxychlor (24.29)
Dicofol (24.29)
24.45
24.62
24.91
25.44
26.19 Mi rex (26.19)
26.52
26.75
27.41
28.07
28.35
29.00
"The GC operating conditions are given in Table 7.
"These are sequentially numbered from elution order and are not isomer numbers
                                          8081  -  36
     Revision  0
September 1994

-------
              TABLE 9  SUMMARY OF RETENTION TIMES  (MIN) OF AROCLORS
                               ON  THE DB-1701  COLUMN"
                              DUAL SYSTEM OF ANALYSIS
Peak Aroclor Aroclor Aroclor
No.b 1016 1221 1232
1 4.45 4.45
2 5.38
3 5.78
4 5.86 5.86
5 6.33 6.31 6.34
6 6.78 6.78 6.79
7 6.96 6.96 6.96
8 7.64
9 8.23 8.23 8.23
10 8.62 8.63 8.63
11 8.88 8.89
12 9.05 9.06 9.06
13 9.46 9.47
14 9.77 9.79 9.78
15 10.27 10.29 10.29
16 10.64 10.65 10.66
17
18 11.01 11.02
19 11.09 11.10
20 11.98 11.99
21 12.39 12.39
22 12.77
23 12.92
24 12.99 13.00
25 13.14 13.16
26
27 13.49
28 13.58
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
13.49
13.61

14.08
14.30

14.49


15.38
15.65
15.78
16.13




16.77
17.13









Aroclor
1242
6.28
6.72
6.90
7.59
8.15
8.57
8.83
8.99
9.40
9.71
10.21
10.59
10.96
11.02
11.94
12.33
12.71
12.94
13.09
13.44
13.54
13.67
14.03
14.26

14.46


15.33
15.62
15.74
16.10




16.73
17.09

17.46
17.69

18.48


19.13

Aroclor
1248
6.91
8.16
8.83
8.99
9.41
9.71
10.21
10.59
10.95
11.03
11.93
12.33
12.69
12.93
13.09
13.44
13.54

14.03
14.24
14.39
14.46

15.10
15.32
15.62
15.74
16.10




16.74
17.07

17.44
17.69
18.19
18.49


19.13

Aroclor
1254
10.95
11.93
12.33
13.10
13.24

13.51
13.68
14.03
14.24
14.36

14.56
15.10
15.32
15.61
15.74
16.08

16.34
16.44
16.55
16.77
17.07
17.29
17.43
17.68
18.17
18.42
18.59
18.86
19.10
19.42
Aroclor Pesticide eluting at same
1260 retention time
Trifluralin (6.96)

13.52

14.02
14.25


14.56

Chlordane (15.32)
16.61 4,4'-DDE (15.67)
15.79

16.19
16.34
16.45

16.77 Perthane (16.71)
17.08
17.31
17.43
17.68
18.18
18.40

18.86
19.09 Endosutfan II (19.05)
19.43
"The GC operating conditions are given in Table 7.
                                                                            (continued)
                                     8081  -  37
    Revision 0
September 1994

-------
                                     TABLE  9   CONTINUED
Peak
No.
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
Aroclor Aroclor Aroclor Aroclor Aroclor Aroclor
1016 1221 1232 1242 1248 1254
19.55
20.20
20.34

20.57 20.55
20.62
20.88

21.53
21.83
23.31








Aroclor Pesticide eluting at same
1260 retention time
19.59 4.4'-DDT (19.54)
20.21

20.43

20.66 Endrin aldehyde (20.69)
20.87
21.03
21.53
21.81
23.27
23.85
24.11
24.46
24.59
24.87
25.85
27.05
27.72
"The GC operating  conditions are given in Table 7.
"These are sequentially numbered from elution order and are not isomer numbers
                                          8081 -  38
     Revision  0
September 1994

-------
                                   TABLE 10
            PEAKS DIAGNOSTIC OF PCBs OBSERVED IN 0.53 mm ID COLUMN
                            SINGLE  COLUMN ANALYSIS


Peak         RT on      RT  on                            Elution
 No.c       DB 608s   DB 1701"  Aroclor"                  Order

I479047661221Before TCmX

II           7.15        6.96   1221, 1232,  1248         Before a-BHC

III          7.89        7.65   1061, 1221,  1232, 1242,  Before a-BHC

IV           9.38        9.00   1016, 1232,  1242, 1248,  just after a-BHC  on
                                                        DB-1701;just before
                                                        7-BHC on  DB-608

V           10.69       10.54   1016. 1232,  1242.        1248 a-BHC and
                                                        heptachlor on DB-1701;
                                                        just after heptachlor
                                                        on DB-608

VI          14.24       14.12   1248, 1254               7-BHC and heptachlor
                                                        epoxide on DB-1701;
                                                        heptachlor epoxide and
                                                        7-Chlordane on DB-608

VII         14.81       14.77   1254                     Heptachlor epoxide and
                                                        7-Chlordane on
                                                        DB-1701; a- and
                                                        7-Chlordane on DB-608

VIII        16.71       16.38   1254                     DDE and Dieldrin on
                                                        DB-1701; Dieldrin and
                                                        Endrin on DB-608

IX          19.27       18.95   1254, 1260               Endosulfan II on
                                                        DB-1701; DDT on DB-608
                                                       Continued
                                  8081 - 39                         Revision  0
                                                                September 1994

-------
                             TABLE 10 (Continued)
            PEAKS DIAGNOSTIC OF PCBs OBSERVED IN 0.53 mm ID COLUMN
                            SINGLE COLUMN  ANALYSIS
Peak         RT on      RT  on                            Elution
 No.        DB 60S8   DB 1701"     Aroclorb              Order
X           21.22       21.23       1260                 Endrin  aldehyde  and
                                                        Endosulfan  sulfate on
                                                        DB-1701;  Endosulfan
                                                        sulfate and
                                                        Methoxychlor on
                                                        on  DB-608

XI          22.89       22.46       1260                 Just  before endrin
                                                        ketone  on DB-1701;
                                                        after endrin ketone  on
                                                        DB-608
0  Temperature program:  Tj =  150°C, hold 30 seconds; increase temperature at
   5°C/minutes to 275°C.

b  Underlined Aroclor indicates the largest peak in the pattern.

c  These are sequentially numbered from elution order and are not isotner
   numbers
                                   8081  -  40                         Revision 0
                                                                September 1994

-------
                 TABLE 11  SPECIFIC PCB CONGENERS IN AROCLORS
Congener
IUPAC number
                  Aroclor
1016  1221  1232  1242  1248  1254  1260
Biphenyl
2CB
23DCB
34DC8
244'TCB
22'35'TCB
23'44'TCB
233'4'6PCB
23'44'5PCB
22'44'55'HCB

1
5
12
28*
44
66*
110
118*
153
X
XXX
XXX
X X
X X
X





X
X
X
X
X






X
X
X
X
X







X
X
X
X
X







X

X
X
22'344'5'HCB         138
22'344'55'HpCB       180
22'33'44'5HpCB       170
                                                      X
                                                      X
                                                      X
*apparent co-elution of two major peaks:

      28 with 31 {2,4',5 trichloro)
      66 with 95 (2,2',3,5',6 pentachloro)
      118 with 149 (2,2',3,4',5',6 hexachloro)
                                  8081 - 41
                                                  Revision  0
                                              September  1994

-------
                 TABLE  12  ANALYTE RECOVERY  FROM  SEWAGE  SLUDGE
Compound
Sonication
Soxhlet

Hexachloroethane
2-Chloronapthalene
4-Bromodiphenyl ether
a-BHC
7-BHC
Heptachlor
Aldrin
0-BHC

-------
               TABLE  13   ANALYTE  RECOVERY  FROM DCE  STILL BOTTOMS
Compound
Sonication
Soxhlet

Hexachloroethane
2-Chloronapthalene
4-Bromodiphenyl ether
a-BHC
0-BHC
Heptachlor
Aldrin
j8-BHC
5-BHC
Heptachlor epoxide
Endosulfan I
7-Chlordane
a-Chlordane
DDE
Dieldrin
Endrin
Endosulfan II
DDT
Endrin aldehyde
ODD
Tetrachloro-m-xylene
Decachl orobi phenyl
%Recovery
70
59
159
55
43
48
48
51
43
47
47
48
45
45
45
50
49
49
40
48
49
17
%RSD
2
3
14
7
6
6
5
7
4
6
4
5
5
4
5
6
5
4
4
5
2
29
%Recovery
50
35
128
47
30
55
200
75
119
66
41
47
37
70
58
41
46
40
29
35
176
104
%RSD
30
35
137
25
30
18
258
42
129
34
18
13
21
40
24
23
17
29
20
21
211
93
Concentration spiked in the sample:  500-1000 ng/g
Three replicates/sample

Extraction solvent, Method 3540 - methylene chloride
Extraction solvent, Method 3550 - methylene chloride/acetone (1:1)

Cleanup - Method 3640

GC column - DB-608, 30M X 0.53 mm ID
                                  8081  - 43
                                   Revision 0
                               September 1994

-------
                                   TABLE 14
             SINGLE  LABORATORY ACCURACY  DATA  FOR  THE  EXTRACTION OF
        ORGANOCHLORINE PESTICIDES FROM SPIKED CLAY SOIL BY METHOD 3541
                             (AUTOMATED SOXHLET)8
Compound Name                 Spike Level       	% Recovery	
                                                      DB-5              DB-1701
a-BHC
/3-BHC
Heptachlor
Aldrin
Heptachlor epoxide
trans-Chlordane
Endosulfan I
Dieldrin
Endrin
Endosulfan II
4, 4' -DDT
Mirex
500
500
500
500
500
500
500
500
500
500
500
500
89
86
94
b
97
94
92
b
111
104
b
108
94
b
95
92
97
95
92
113
104
104
b
102
a     The  operating  conditions  for  the  automated Soxhlet  were  as  follows:
      immersion time 45 min; extraction time  45  min;  the  sample size was 10 g
      clay soil,  extraction solvent,  1:1 acetone/hexane.  No equilibration time
      following spiking.

b     Not able to determine because of interference.

Data taken from Reference 14.
                                   8081  -  44                         Revision 0
                                                                September 1994

-------
                         TABLE  15
     SINGLE LABORATORY RECOVERY DATA FOR EXTRACTION OF
PCBS FROM CLAY AND SOIL BY METHOD 3541" (AUTOMATED SOXHLET)
Matrix Compound Spike Level
(ppm)
Clay Aroclor-1254 5





Clay Aroclor-1254 50





Clay Aroclor-1260 5





Clay Aroclor-1260 50





Soil Aroclor-1254 5




Soil Aroclor-1254 50





Trial
1
2
3
4
5
6
1
2
3
4
5
6
1
2
3
4
5
6
1
2
3
4
5
6
1
2
3
4
5
1
2
3
4
5
6
Percent
Recovery6
87.0
92.7
93.8
98.6
79.4
28.3
65.3
72.6
97.2
79.6
49.8
59.1
87.3
74.6
60.8
93.8
96.9
113.1
73.5
70.1
92.4
88.9
90.2
67.3
69.7
89.1
91.8
83.2
62.5
84.0
77.5
91.8
66.5
82.3
61.6
                                             (continued)
                         8081 - 45
    Revision 0
September 1994

-------
                                   TABLE 15
                                  (continued)
Matrix Compound Spike Level
(ppm)
Soil Aroclor-1260 5






Soil Aroclor-1260 50





Trial
1
2
3
4
5
6
7
1
2
3
4
5
6
Percent
Recovery6
83.9
82.8
81.6
96.2
93.7
93.8
97.5
76.9
69.4
92.6
81.6
83.1
76.0
a     The  operating  conditions  for the  automated Soxhlet  were  as  follows:
      immersion time 60 min; reflux time 60 min.

b     Multiple results from two different extractors.

Data from Reference 15.
                                  8081  - 46
    Revision 0
September 1994

-------
           TABLE  16.  MULTI-LABORATORY  PRECISION AND ACCURACY  DATA
                  FOR THE EXTRACTION OF  PCBS  FROM SPIKED  SOIL
                      BY METHOD 3541  (AUTOMATED SOXHLET)
	
Laboratory
Lab 1
Lab 2
	
Lab 3
Lab 4
Lab 5
	 . .
Lab 6
Lab 7
Lab 8
All
Laboratories
	
Num
Average
St Dev
Num
Average
St Dev
Num
Average
St Dev
Num
Average
St Dev
Num
Average
St Dev
	
Num
Average
St Dev
Num
Average
St Dev
	
Num
Average
St Dev
Num
Average
St Dev
	 	 • 	
PCB Percent Recovery
Aroclor
	 12.54 	
PCB Level
0
3.0
101.2
34.9

	
3.0
72.8
10.8
6.0
112.6
18.2

	
2.0
140.9
4.3
3.0
100.1
17.9
3.0
65.0
16.0
	
20.0
98.8
28.7
50
	
3.0
74.0
41.8
6.0
56.5
7.0
3.0
63.3
8.3
6.0
144.3
30.4
3.0
97.1
8.7
3.0
127.7
15.5
3.0
123.4
14.6
3.0
38.3
21.9
	
30.0
92.5
42.9
500

6.0
66.9
15.4


3.0
80.1
5.1



9.0
71.3
14.1
1260
PCB Level
5
3.0
83.9
7.4

	
3.0
70.6
2.5
6.0
100.3
13.3

....
3.0
138.7
15.5
3.0
82.1
7.9
3.0
92.8
36.5
21.0
95.5
25.3
	
50
3.0
78.5
7.4
6.0
70.1
14.5
3.0
57.2
5.6
6.0
84.8
3.8
3.0
79.5
3.1
4.0
105.9
7.9
3.0
94.1
5.2
3.0
51.9
12.8
	
31.0
78.6
18.0
500

6.0
74.5
10.3

	
3.0
77.0
9 4
	 .r. :.".
	
	
	
9.0
75.3
9.5
All
Levels
12.0
84.4
26.0
24.0
67.0
13.3
12.0
66.0
9.1
24.0
110.5
28.5
12.0
83.5
10.3
12.0
125.4
18.4
12.0
99.9
19.0
12.0
62.0
29.1
120.0
87.6
29.7
Data from Reference 13.
                                  8081  - 47
    Revision 0
September 1994

-------
                                       FIGURE 1.
         GAS  CHROMATOGRAM  OF THE MIXED  ORGANOCHLORINE PESTICIDE STANDARD
        Surt I'W : 0.00 min
        Scale factor:  0
End Tint  : 33.00 Kin
Plot Off»*t: 20 n«
                    Low Paine : 20.00 nV
                    Plat Sc«U: 100 UN
                                                                Hijn Point : 420.00 M
                                       Response  [mV]
       o-

     :o
     j
      I  I I I  I I I  I I I  I I
                       =-7 . 99
                                             9.93
                                                           -23.18
                             23.80
                              26.23
                            •=-28.64
Column:
Temperature program:
                                                                            —0.95
                                                                            -8.60
                                             — 30.19
30 m x  0.25 mm ID,  DB-5
100°C  (hold 2  minutes)  to  160°C  at  15°C/min,  then at
5°C/min to  270°C;  carrier He  at 16 psi.
                                     8081  -  48
                                                Revision 0
                                           September 1994

-------
                                       FIGURE 2.

     GAS  CHROMATOGRAM  OF INDIVIDUAL ORGANOCHLORINE  PESTICIDE  STANDARD MIX  A


          Start Time : 0.00 min      End lime  : 53.00 *in       Lou Point : 20.00 «V      mgn Point : 270.CO mv
          Scale factor:  0         Plot offset: 20 .w         Plot Scale: 250 mv
                                          Response  [mV]
         o-
                              LP
                              o
                                   o
                                   o
(Ji
o
O
O
                        I  I  I   I  I  I  I  I   I  I  I  I   I  I  I  I  I
        ro
        (_n
        O

I  |  I   I  I  I  I
2.

D

o'


H

3'
0)
         to
          -
                           •7.93
                             I. 60
                           -14.27
                            -17.08
                                 0.22
                                 1.77
                                   22.68
                                 -23.73
                                  --28.52
                                                                               -4.95

-12.33


                                                          -9.86
                                                                    -17.54
                                                                     -18.47
                                          -19.78
                                                             -19.24
                                                        -21.13
                                                                               -23.03
                                                                         -30.05
Column:
Temperature program:
                    30 m  x  0.25 mm  ID,  DB-5
                    100°C  (hold  2  minutes) to  160°C  at 15°C/min,  then  at
                    5°C/min  to  270°C; carrier  He at  16 psi.
                                      8081  -  49
                                                                    Revision 0
                                                                September 1994

-------
                                  FIGURE 3

GAS CHROMATOGRAM  OF INDIVIDUAL  ORGANOCHLORINE PESTICIDE STANDARD MIX  B
   Start rifne : 0.00 mm
   Scale Factor:  0
                           End Ti«  :  JJ.OO "in

                           Plot Offset:  20 ">V
Low Point : 20.00 mv
Plot Scale: 250 mV
                                                               ?:r". : 270.?C mv
                                  Response [mV]
  o-
                       o
                       I   i  i  i  i
                                       -i          -»          K)
                                       O          Ui          O
                                       000

                                       I   I  I  I  I  I  I   I  I  I  I
                         to
                         (_n
                         O
                 I'll
    0)


    3


    o' ->_!
    3
    CD
                   •2.74
                    -6.97
                  tL-9.60
                              --10.71
                    -14.27
                       . 24
                      ^20.69
                         22.00
                                        -11.73
                                            -14.84
                                       -16.23
                                        —17.OS
                                        -17.63
                                                 -18.31


                                                 	19.54
                                             -20.19

                                             	21.03
                                                   --22.68
                                                                        -4.95
                                                              -30.04
Column:
Temperature program:
                     30  m  x 0.25 mm  ID,  DB-5
                     100°C  (hold  2  minutes) to  160°C  at 15°C/min,  then  at
                     50C/min  to 270°C; carrier  He at  16 psi.
                                 8081 - 50
                                                                          Revision 0
                                                                      September 1994

-------
                                       FIGURE 4.
                     GAS CHROMATOGRAM OF THE TOXAPHENE STANDARD
      Start Hint : 0.00 mm
      Sole Factor:  0
End lime  : 33.00 nin
Plot Offset: 20 i«v
low Point : 20.00 mV
Plot Scale: 60 inv
                                                              KijH Point : 80.00 mv
                                     Response [mV]
                   K>        (_J        -f*        <_fl        O>        -J
                   o        o        o        o        o        o
                    lii ill 11 ii 11 ii 1111 ill it H lii ill 1111 lini 111 ill mi 111111 mi
  §SH
  'I'
  1  ' ro
  z!  °"
     KJ
     Ul"
     O
                            ',99
                                                                            24 . 32
Column:
Temperature program:
 30  m x 0.25 mm ID, DB-5
 100°C  (hold  2 minutes)  to 160°C  at  15°C/min,  then  at
 50C/min  to 270°C;  carrier He  at  16 psi.
                                      8081 -  51
                                                  Revision 0
                                             September 1994

-------
                                       FIGURE  5.
                   GAS  CHROMATOGRAM OF THE AROCLOR-1016  STANDARD

         Start lime : 0.00 min      End lime  : 33.00 min       Low Point : 20.00 Hi      "<9K Point : UO.OO mv
         Scale Factor:  0         Plot Offset: 20 mv         Plot Sole: 100 mv
                                        Response  [mV]
        o-
                      N>         ^         O)          CO          O
                      O         O         O          O          O
                       I 1111 111  I) 11 111 1111  111 1111 1111 111  1111 111 1111  111 111
        on—
        O~
       ' f-o
         "
        L-1"
        '-••I
                                  -1.81
                                         9. 83
                                                       	12.95
                                                                             -1.03
Column:
Temperature program:
30 m x  0.25 mm  ID  DB-5 fused  silica capillary.
100°C  (hold 2  minutes)  to  160°C  at  15°C/min,  then at
5°C/min to  270°C;  carrier  He at 16 psi.
                                     8081  - 52
                                                Revision 0
                                           September 1994

-------
                                  FIGURE  6.
           GAS CHROMATOGRAM OF  THE TECHNICAL  CHLORDANE  STANDARD

    Start rime  : 0.00 mm      End lime  : J3.00 Bin      Low Point : 20.00 mV      High Point : 220.00 off
    Scale Factor;  0        Plot Offset: 20 mv         Plot Sc«lt: 200 nv
                                   Response  [mV]
                         mow
                         o            a            o
                  i  i   j   I  i   i   i   i  I   i   i  i   i   i
                                                                     tO
                                                                     o
                                                                     o
                                                              i  i   i   I  i   i
   O"
.11
ftl
   LH"
                                4.59
                      4.33


                   =-5.83
                         -8.87
                                           13.60
                                    38
                                                                        -0.97
                                                           17. 11
                                                          17.65
Column:
Temperature program:
                       30 m x 0.25  mm ID DB-5  fused silica  capillary.
                       100°C  (hold  2  minutes)  to  160°C  at  15°C/min,  then  at
                       5°C/min  to 270°C;  carrier He  at  16 psi.
                                  8081  -  53
                                                                          Revision 0
                                                                     September 1994

-------
                                                       r-
   Ll
          OB-1701
          DB-5
FIGURE 7.    GC/ECD chromatogram of Toxaphene analyzed on a DB-5/DB-1701 fused-
            silica open-tubular column pair.  The GC operating conditions were
            as follows:  30 m x 0.53 mm ID DB-5 (1.5-jum film thickness) and 30
            m x 0.53 mm  ID DB-1701  (1.0-jun film thickness) connected to a J&W
            Scientific press-fit Y-shaped inlet  splitter.   Temperature program:
            150°C  (0.5  min  hold)  to  190°C (2 min hold) at 12°C/min then  to  275°C
            (10 min hold) at 4°C/mn.
                                   8081  -  54
    Revision 0
September 1994

-------
                                                          0
              DB-1701
              0-
              r-
              DB-5
      !•»
      «•
      tfl
        &
    •14  t lit
    um  a- s> — AJ . .
     M  ru   i>«

     I-AT
                                                                     >
FIGURE 8.   GC/ECD chromatogram  of Strobane analyzed on  a  DB-5/DB-1701  fused-
            silica open-tubular  column  pair.   The  GC operating conditions were
            as follows:  30 m x  0.53 mm  ID  DB-5  {1.5-jum film thickness)  and 30
            m x 0.53 mm  ID  DB-1701 (1.0-/im film thickness)  connected  to a J&W
            Scientific press-fit Y-shaped inlet splitter.  Temperature program:
            150°C  (0.5 min  hold)  to 190°C (2 min  hold) at  12°C/min then to 275°C
            (10 min hold) at 4°C/niin.
                                   8081 - 55
    Revision 0
September 1994

-------
       If
       0
                         ra
                   10-O





> f-

                                            o
                                           lUl'4
                                           a* •
                                           r- u
FIGURE 9.   GC/ECD  chromatogram  of  Aroclor  1016  analyzed  on  a  DB-5/DB-1701
            fused-silica open-tubular column pair.  The GC operating conditions
            were as follows:  30 m x 0.53 mm ID DB-5 (1.5-jum film thickness) and
            30 m x 0.53 mm  ID DB-1701 (1.0-^m film thickness) connected to a J&W
            Scientific press-fit Y-shaped inlet splitter.  Temperature program:
            150°C (0.5 min  hold) to 190°C  (2 min  hold)  at  12°C/niin  then to 275°C

            (10 min hold)  at 40C/min.
                                   8081 - 56
    Revision 0
September 1994

-------
                 r-to .
                 f^O- *•
                                                       DB-1701
                           Or-
                           C-0
                            •CO
                            'v/v

                                                       OB-5
FIGURE 10.  GC/ECD  chromatogram of  Aroclor  1221  analyzed  on a  DB-5/DB-1701
            fused-silica open-tubular column pair.  The GC operating conditions
            were as follows:  30 m x 0.53 mm ID DB-5 (1.5-/im film thickness) and
            30 m x 0.53 mm  ID DB-1701 (l.Q-pm  film thickness) connected to a J&W
            Scientific press-fit Y-shaped inlet splitter.  Temperature program:
            150°C (0.5 min  hold) to  190°C  (2 min hold)  at  12°C/min then to 275°C
            (10 min hold) at 4°C/nrin.
                                   808] - 57
    Revision 0
September 1994

-------
                                                   DB-1701
      r
                                                   OB-5
FIGURE 11.  GC/ECD  chromatogram  of  Aroclor  1232  analyzed  on  a DB-5/D8-1701
            fused-silica open-tubular column  pair.  The GC operating conditions
            were as follows:  30 m x 0.53 mm ID  DB-5 (1.5-jum  film thickness) and
            30 m x 0.53 mm ID DB-1701 (1.0-jum film thickness) connected to a J&W
            Scientific press-fit  Y-shaped inlet splitter.  Temperature program:
            150°C (0.5 min hold) to  190°C (2 min hold) at 12°C/min then to 275°C
            (10 min hold) at 4°C/min.
                                   8081  -  58
    Revision 0
September 1994

-------

-------
           0
           (u
                           in
                           tn
                         r«
                     OD-bu tl)
              -111
           IJL
   I
 m t




uW
                                                               DB-1701
                                                                  r\i
                                                                  rv
                                                         ID  K)
                                                              DB-5
                a)  tr.
                C.  M
FIGURE 13.  GC/ECD  chromatogram  of Aroclor  1248  analyzed  on  a DB-5/DB-1701
            fused-silica open-tubular column  pair.  The GC operating conditions
            were as follows:  30 m x 0.53 mm ID DB-5 (1.5-/itn  film  thickness)  and
            30 m x 0.53 mm  ID DB-1701 (1.0-^m  film thickness) connected to a  J&W
            Scientific press-fit  Y-shaped  inlet splitter.  Temperature  program:
            150°C (0.5 min  hold) to  190°C (2 min hold) at 12°C/min then to 275°C
            (10 min hold)  at 4°C/min.
                                   8081 - 60
                                           Revision 0
                                       September 1994

-------
                                          CD
                                     in

                                     m
                                                             DB-1701
       u
                                                             DB-5
         0 C«'l  u Ui M
FIGURE 14.  GC/ECD  chromatogram of  Aroclor  1254  analyzed  on  a  DB-5/OB-1701
            fused-silica open-tubular column pair.   The GC operating conditions
            were as follows:   30 m x 0.53 mm ID DB-5 (1.5-jum film thickness) and
            30 m x 0.53 mm  ID D8-1701 (1.0-jum  film thickness) connected to a J&W
            Scientific press-fit Y-shaped inlet splitter.  Temperature program:
            150°C (0.5 min  hold) to  190°C  (2 min hold)  at  12°C/nnn then to 275°C
            (10 min hold) at 4°C/min.
                                   8081 -  61
    Revision 0
September 1994

-------
                         DB-1701
                        DB-5
FIGURE 15.   GC/ECD  chromatogram  of Aroclor  1260  analyzed  on  a  DB-5/DB-1701
            fused-silica open-tubular  column  pair.  The GC operating conditions
            were as  follows:  30 m x 0.53 mm ID DB-5 (1.5-^m film thickness) and
            30 m x 0.53 mm ID DB-1701 (1.0-/*m  film thickness)  connected  to a J&W
            Scientific press-fit  Y-shaped inlet  splitter.  Temperature program:
            150°C (0.5 min hold)  to  190°C (2 min hold) at 12°C/min  then  to 275°C
            (10 min  hold) at 4°C/imn.
                                  8081  - 62
    Revision 0
September 1994

-------
                                                         DB-1701
         orvi r- —
         (mrr o -
         OOT'IO T
                Q-T
                       CD
                     y irrn
                                                           0>
                                                           o
                                                                   K> rn
                                                                   n  ®
                                                                   — 01
 r-
 01
                                                        DB-5
FIGURE 16.  GC/ECD  chromatogram  of Halowax  1000  analyzed  on  a DB-5/DB-1701
            fused-silica open-tubular column  pair.  The GC operating conditions
            were as follows:  30 m x 0.53  mm ID DB-5 (1.5-jum  film thickness)  and
            30 m x 0.53 mm  ID DB-1701 (1.0-/im  film thickness) connected to a  J&W
            Scientific press-fit  Y-shaped inlet splitter.  Temperature program:
            150°C (0.5 min hold)  to  190°C (2 min hold) at 12°C/nun then to 275°C
            (10 min hold)  at 4°C/min.
                                   8081 - 63
    Revision 0
September 1994

-------
                                                08-1701
FIGURE 17.   GC/ECD  chromatogram of  Halowax 1001  analyzed on  a DB-5/DB-1701
            fused-silica open-tubular column pair.   The GC  operating conditions
            were as follows:   30 m x 0.53 mm ID  DB-5 (1.5-/nm film thickness)  and
            30 m x 0.53 mm ID DB-1701 (1.0-jum film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature  program:
            150°C (0.5 min hold) to  190°C (2 min hold) at 12°C/nnn then to 275°C
            (10 min hold) at 4°C/nnn.
                                   8081  -  64
    Revision 0
September 1994

-------
                                                          OB-1701
                             •a
                             »
9-

II
                                                          DB-5
                  •o
FIGURE 18.  GC/ECD  chromatogram of  Halowax.1099  analyzed  on  a  DB-5/DB-1701
            fused-silica open-tubular column pair.   The GC  operating  conditions
            were as follows:   30 m x 0.53 mm ID DB-5 (1.5-/um film thickness)  and
            30 m x 0.53 mm ID DB-1701 (1.0-^m film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature program:
            150°C (0.5 min hold) to  190°C  (2 min hold)  at 12°C/min then to 275°C
            (10 min hold) at 4°C/niin.
                                   8081  -  65
                               Revision 0
                           September 1994

-------
                   u — K
                   _ |. t«
                    • 0 »
                                                             DB-1701
                                                                •0
                                                                0
                                                          e
                                                          o
                                                            DB-5
                  HI

                  •t
FIGURE 19.  GC/ECD  chromatogram of  Halowax  1013  analyzed on  a DB-5/DB-1701
            fused-silica open-tubular column  pair.   The GC  operating conditions
            were as follows:   30 m x 0.53 mm ID  DB-5 (LS-^m film thickness)  and
            30 m x 0.53 mm ID DB-1701 (1.0-fim film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature  program:
            150°C (0.5 min hold) to  190°C (2 min hold) at 120C/min then to 275°C
            (10 min hold) at 40C/min.
                                   8081  -  66
    Revision 0
September 1994

-------
                        DB-1701
                         . .1
                   LT^T -. -. %
                  r-fl -.-" " *>
FIGURE 20.  GC/ECD  chromatogram of  Halowax  1014  analyzed  on  a  DB-5/DB-1701
            fused-silica open-tubular column'pair.   The GC operating conditions
            were as follows:   30 m x 0.53 mm ID DB-5 (1.5-^m film thickness) and
            30 m x 0.53 mm  ID DB-1701 (1.0-^cm  film thickness) connected to a J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature program:
            150°C (0.5 min  hold) to  190°C  (2 min hold)  at  12°C/min then to 275°C
            (10 min hold) at 4°C/min.
                                   8081 -  67
    Revision 0
September 1994

-------
                            DB-1701
                                       •» I • * 9W
                       J I"" -•."••
                                               "••» •• - f
                                               — — -— "•« t* f
                             DB-5
                     i
                       Ju, : : ;f
                       liii
FIGURE 21.  GC/ECD  chromatogram  of  Halowax  1051  analyzed  on  a  DB-5/DB-1701
            fused-silica open-tubular column  pair.  The GC operating conditions
            were as follows:  30 m x 0.53 mm ID  DB-5 (1.5-jum  film thickness) and
            30 m x 0.53 mm ID DB-1701 (1.0-/im film thickness) connected to a J&W
            Scientific press-fit  Y-shaped inlet splitter.  Temperature program:
            150°C (0.5 min hold) to  190°C (2 min hold) at 12°C/nnn  then to 275°C
            (10 min hold) at 4°C/min.
                                   8081  -  68
    Revision 0
September 1994

-------
              IS SU
                                      DB-5
                                    a
                                           24
2»
34 33    42
                                              '14
                                                      37
                                                      3S
                                                         43
                                     DB-1701
     12      1    4 SU IS   (
                               10  II 12  It
                                        14
                                                          32
                                                             4 34   )»   4
                                                                        43
                                                                              20.
FIGURE 22.  GC/ECD  chromatogram of the organochlorine pesticides  analyzed on a
            DB-5/DB-1701   fused-silica  open-tubular  column   pair.     The  GC
            operating conditions were  as  follows:  30 m x 0.53 mm ID DB-5 (0.83-
            jjm  film thickness) and 30  m  x  0.53 mm  ID DB-1701  (1.0-/im  film
            thickness)  connected  to  an  8  in  injection tee  (Supelco  Inc.).
            Temperature  program:   140°C  (2 min  hold)  to 270°C (1 min  hold)  at
            2.8°C/Tiin.
                                   8081  -  69
                      Revision 0
                  September 1994

-------
                                       METHOD 8081

               ORGANOCHLORINE  PESTICIDES AND PCBs AS  AROCLORS BY  GAS
                     CHROMATOGRAPHY:  CAPILLARY  COLUMN TECHNIQUE
     7.1.1 ChooM
  appropriate extraction
technique (sea Ctiaptar 2)
   7.1. 2 Add specified
 matlx spfca to sample.
  7.2 Routine cleanup/
     fractfonation.
7.3 Set chromatograpNc
      cendHtens.
         i
7.4 Refer lo Method 8000
  tor proper caJbratfon
     techniques.
                                 7.4.2 Prime or deactivate GC
                                  column prior to caHbraVon.
                                           E
                                 7.5 Perform GC analysis (s
                                       MettKxJSOOO)
                                                                        7.5.8 AddtkanaJ
                                                                     cteanuprtracttonatton
                                                                       (saeSectfon7.2)
                                                                       7.6 Calculation of
                                                                  toxapheoe, crtoroane, PCBs.
                                                                    DDT, and BHC done here.
                                       8081 -  70
                                                                               Revision  0
                                                                           September 1994

-------
                                LIST OF TABLES

Table 1     Gas chromatographic retention times and  method detection limits for
            the Organochlorine Pesticides and PCBs as Aroclors using wide-bore
            capillary columns, single column analysis

Table 2     Gas chromatographic retention times and  method detection limits for
            the Organochlorine pesticides and PCBs as Aroclors  using narrow-bore
            capillary columns, single column analysis

Table 3     Estimated quantitation limits (EQL) for various matrices

Table 4     GC Operating conditions for  Organochlorine compounds, single column
            analysis

Table 5     Retention times of the Organochlorine pesticides, dual column method
            of analysis

Table 6     GC operating conditions for Organochlorine pesticides,  dual column
            method of analysis, low temperature, thin film

Table 7     GC operating conditions for Organochlorine pesticides,  dual column
            method of analysis, high temperature, thick film

Table 8     Summary of  retention  times  (min)  of Aroclors on  the DB  5 column,
            dual system of analysis

Table 9     Summary of retention times (min)  of Aroclors on the DB 1701 column,
            dual system of analysis

Table 10    Peaks  diagnostic  of PCBs observed in  0.53  mm ID  column,  single
            column system of analysis

Table 11    Specific Congeners in Aroclors

Table 12    Recovery from Sewage Sludge

Table 13    Recovery DCE still bottoms

Table 14    Single Laboratory Accuracy Data for the Extraction  of Organochlorine
            Pesticides from Spiked Clay  Soil  by Method 3541 (Automated Soxhlet)

Table 15    Single Laboratory Recovery Data for Extraction of PCBs from Clay and
            Soil by Method 3541 (Automated Soxhlet)

Table 16    Multi-laboratory Precision and Accuracy Data for the Extraction of
            PCBs from Spiked Soil by Method 3541 (Automated Soxhlet)
                                   8081  -  71
    Revision 0
September 1994

-------
                                LIST OF FIGURES

Figure 1.   GC of the Mixed Organochlorine Pesticide Standard.  The GC operating
            conditions  were  as  follows:   30  m x  0.25  mm  ID  DB-5  column.
            Temperature program:   100°C  (hold  2  minutes)  to 160°C at 15°C/min,
            then at 5°C/min to 270°C;  carrier  He at  16 psi.

Figure 2.   GC of  Individual  Organochlorine  Pesticide Standard Mix A.   The  GC
            operating  conditions  were  as  follows:    30 m x  0.25  mm  ID  DB-5
            column.   Temperature  program:   100°C (hold 2  minutes)  to 160°C  at
            15°C/min,  then at 5°C/nnn  to 270°C;  carrier  He  at 16  psi.

Figure 3.   GC of  Individual  Organochlorine  Pesticide Standard Mix B.   The  GC
            operating  conditions  were  as  follows:    30 m x  0.25  mm  ID  DB-5
            column.   Temperature  program:   100°C (hold 2  minutes)  to 160°C  at
            15°C/min,  then at 5°C/min  to 270°C;  carrier  He  at 16  psi.

Figure 4.   GC of the Toxaphene Standard.  The GC operating conditions  were  as
            follows:   30 m  x  0.25 mm  ID  DB-5  column.   Temperature program:
            100°C (hold 2 minutes) to 160°C at 15°C/min> then at 5°C/min to 270°C;
            carrier He  at 16 psi.

Figure 5.   GC of the Aroclor-1016  Standard.  The GC  operating conditions  were
            as follows:  30 m x 0.25 mm  ID .DB-5  fused silica capillary  column.
            Temperature program:   100°C  (hold  2  minutes)  to 160°C at 15°C/min,
            then at 5°C/min to 270°C;  carrier  He at  16 psi.

Figure 6.   GC of the Technical  Chlordane Standard.  The GC operating conditions
            were as  follows:   30 m x 0.25 mm  ID DB-5  fused  silica  capillary
            column.   Temperature  program:   100°C (hold 2  minutes)  to 160°C  at
            15°C/min,  then at 5°C/rnin  to 270°C;  carrier  He  at 16  psi.

Figure 7.   GC/ECD chromatogram of  Toxaphene analyzed on  a DB-5/DB-1701 fused-
            silica open-tubular column  pair.  The GC  operating conditions  were
            as follows:  30 m x 0.53 mm  ID DB-5  {1.5-jum film thickness) and  30
            m x  0.53  mm ID DB-1701  (1.0-/xm film thickness)  connected to a J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature program:
            150°C (0.5 min hold) to 190°C (2 min  hold) at 12°C/min then to 275°C
            (10 min hold) at 4°C/min.

Figure 8.   GC/ECD chromatogram of  Strobane  analyzed on a  DB-5/DB-1701 fused-
            silica open-tubular column  pair.  The GC  operating conditions  were
            as follows:  30 m x 0.53 mm  ID DB-5  (1.5-jum film thickness) and  30
            m x  0.53  mm ID DB-1701  (1.0-jum film thickness)  connected to a J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature program:
            150°C (0.5 min hold)  to 190°C (2 min  hold) at 12°C/nrin then to 275°C
            (10 min hold) at 4°C/min.
                                   8081 - 72
    Revision 0
September 1994

-------
Figure 9.   GC/ECD  chromatogram of  Aroclor 1016  analyzed  on  a  DB-5/DB-1701
            fused-silica open-tubular column pair.   The GC  operating  conditions
            were as follows:   30 m x 0.53 mm ID DB-5 (1.5-^tm film thickness)  and
            30 m x 0.53 mm  ID DB-1701 (1.0-^m film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature  program:
            150°C (0.5 min  hold) to  190°C  (2 min hold)  at 12°C/min then to 275°C
            {10 min hold) at 4°C/min.

Figure 10.  GC/ECO  chromatogram of  Aroclor 1221  analyzed  on  a  DB-5/DB-1701
            fused-silica open-tubular column pair.   The GC  operating  conditions
            were as follows:   30 m x 0.53 mm ID DB-5 (1.5-/im film thickness)  and
            30 m x 0.53 mm  ID DB-1701 (l-D-^m film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature  program:
            150°C (0.5 min  hold) to  190°C  (2 min hold)  at 12°C/min then to 275°C
            (10 min hold) at 4°C/min.

Figure 11.  GC/ECD  chromatogram of  Aroclor 1232  analyzed  on  a  DB-5/OB-1701
            fused-silica open-tubular column pair.   The GC  operating  conditions
            were as follows:   30 m x 0.53 mm ID DB-5 (1.5-jiim film thickness)  and
            30 m x 0.53 mm  ID DB-1701 (1.0-p film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature  program:
            150°C (0.5 min  hold) to  190°C  (2 min hold)  at 12°C/min then to 275°C
            (10 min hold) at 4°C/min.

Figure 12.  GC/ECD  chromatogram of  Aroclor 1242  analyzed  on  a  DB-5/DB-1701
            fused-silica open-tubular column pair.   The GC  operating  conditions
            were as follows:   30 m x 0.53 mm ID  DB-5 (1.5-/im film thickness)  and
            30 m x 0.53 mm  ID DB-1701 (1.0-^m film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature  program:
            150°C (0.5 min  hold) to  190°C  (2 min hold)  at 12°C/min then to 275°C
            (10 min hold) at 4°C/min.

Figure 13.  GC/ECD  chromatogram of  Aroclor 1248  analyzed  on  a  DB-5/DB-1701
            fused-silica open-tubular column pair.   The GC  operating  conditions
            were as follows:   30 m x 0.53 mm ID DB-5 (1.5-jum film thickness)  and
            30 m x 0.53 mm  ID DB-1701 (1.0-/im film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature  program:
            150°C (0.5 min  hold) to  190°C  (2 min hold)  at 120C/min then to 275°C
            (10 min hold) at 4°C/nnn.

Figure 14.  GC/ECD  chromatogram of  Aroclor 1254  analyzed  on  a  DB-5/D8-1701
            fused-silica open-tubular column pair.   The GC  operating  conditions
            were as follows:   30 m x 0.53 mm ID  DB-5 (1.5-jum film thickness)  and
            30 m x 0.53 mm  ID DB-1701 (1.0-/um film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature  program:
            150°C (0.5 min  hold) to  190°C  (2 min hold)  at 12°C/min then to 275°C
            (10 min hold) at 4°C/imn.
                                   8081  -  73
    Revision 0
September 1994

-------
Figure 15.  GC/ECD  chromatogram of  Aroclor 1260  analyzed on  a DB-5/DB-1701
            fused-silica open-tubular column pair.   The  GC  operating conditions
            were as follows:   30 m x 0.53 mm ID  DB-5 (1.5-/nm film thickness)  and
            30 m x 0.53 mm ID DB-1701 (1.0-Mm film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature  program:
            150°C (0.5 min hold) to  190°C  (2 min hold) at \2°C/mir\ then to 275°C
            (10 min hold) at 4°C/min.

Figure 16.  GC/ECD  chromatogram of  Halowax 1000  analyzed on  a DB-5/DB-1701
            fused-silica open-tubular column pair.   The  GC  operating conditions
            were as follows:   30 m x 0.53 mm ID  DB-5 (1.5-^m film thickness)  and
            30 m x 0.53 mm ID DB-1701 (1.0-/im film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature  program:
            150°C (0.5 min hold) to  190°C  (2 min hold) at 12°C/nun then to 275°C
            (10 min hold) at 4°C/min.

Figure 17.  GC/ECD  chromatogram of, Halowax 1001  analyzed on  a DB-5/DB-1701
            fused-silica open-tubular column pair.   The  GC  operating conditions
            were as follows:   30 m x 0.53 mm ID  DB-5 (1.5-jum film thickness)  and
            30 m x 0.53 mm ID DB-1701 (1.0-jim film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature  program:
            150°C (0.5 min hold) to  190°C  (2 min hold) at 12°C/min then to 275°C
            (10 min hold) at 4°C/min.

Figure 18.  GC/ECD  chromatogram of  Halowax 1099  analyzed on  a DB-5/DB-1701
            fused-silica open-tubular column pair.   The  GC  operating conditions
            were as follows:   30 m x 0.53 mm ID  D6-5 (1.5-/itn film thickness)  and
            30 m x 0.53 mm ID DB-1701 (1.0-/^m film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature  program:
            150°C (0.5 min hold) to  190°C  (2 min hold) at 12°C/min then to 275°C
            (10 min hold) at 4°C/min.

Figure 19.  GC/ECD  chromatogram of  Halowax 1013  analyzed on  a DB-5/DB-1701
            fused-silica open-tubular column pair.   The  GC  operating conditions
            were as follows:   30 m x 0.53 mm ID  DB-5 (1.5-^m film thickness)  and
            30 m x 0.53 mm ID DB-1701 (1.0-/xm film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature  program:
            150°C (0.5 min hold) to  190°C  (2 min hold) at 12°C/min then to 275°C
            (10 min hold) at 4°C/min.

Figure 20.  GC/ECD  chromatogram of  Halowax 1014  analyzed on  a DB-5/DB-1701
            fused-silica open-tubular column pair.   The  GC  operating conditions
            were as follows:   30 m x 0.53 mm ID  DB-5 (1.5-jum film thickness)  and
            30 m x 0.53 mm ID DB-1701 (1.0-^m film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature  program:
            150°C (0.5 min hold) to  190°C  (2 min hold) at 12°C/min then to 275°C
            (10 min hold) at 4°C/min.
                                   8082  -  74
    Revision 0
September 1994

-------
Figure 21.  GC/ECD  chromatogram of  Halowax 1051  analyzed on  a DB-5/DB-1701
            fused-silica open-tubular column pair.   The GC  operating  conditions
            were as follows:   30 m x 0.53 mm ID  DB-5 (1.5-^m film thickness)  and
            30 m x 0.53 mm ID OB-1701 (1.0-/im film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature program:
            150°C (0.5 min hold) to  190°C  (2 min hold)  at 12°C/min then to 275°C
            (10 min hold) at 4°C/min.

Figure 22.  GC/ECD chromatogram of the organochlorine  pesticides  analyzed on  a
            DB-5/DB-1701  fused-silica  open-tubular   column  pair.    The  GC
            operating conditions were as follows:   30 m x 0.53 mm ID DB-5  (0.83-
            jLim  film thickness)  and  30 m x 0.53  mm   ID DB-1701  (1.0-/^m film
            thickness)  connected to  an  8  in   injection   tee  (Supelco  Inc.).
            Temperature program:   140°C  (2  min hold)  to 270°C (1 min  hold) at
            2.8°C/min.
                                   8081 - 75                         Revision  0
                                                                 September  1994

-------
00
>_t
I—*
o

-------
                                 METHOD 8110

                       HALOETHERS BY GAS CHROMATOGRAPHY


1.0   SCOPE AND APPLICATION

      1.1   This method covers  the  determination of certain haloethers.   The
following compounds can be determined by this method:


                                                   Appropriate Technique
Compound Name                  CAS  No.a    3510    3520    3540    3550    3580
Bis(2-chloroethoxy)methane
Bis(2-chloroethyl) ether
Bts(2-chloroisopropyl) ether
4-Bromophenyl phenyl ether
4-Chlorophenyl phenyl ether
111-91-1
111-44-4
108-60-1
101-55-3
7005-72-3
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
   a  Chemical  Abstract Services  Registry  Number.
   X  Greater than 70 percent recovery by  this technique.

      1.2   This  is  a gas  chromatographic  (GC)   method  applicable  to  the
determination  of  the  compounds  listed  above  in  municipal  and  industrial
discharges.  When this method is used to analyze unfamiliar samples for any or
all of the compounds above, compound identifications should be supported by at
least one  additional  qualitative technique.   This method  describes analytical
conditions of a second GC column that can  be used to confirm measurements made
with  the  primary  column.    Method  8270  provides  gas  chromatograph/mass
spectrometer (GC/MS) conditions appropriate for the qualitative and quantitative
confirmation of results for all of the parameters listed above, using the extract
from this method.

      1.3   The method  detection limit  (MDL,  defined  in  Section  9.1)  for each
parameter  is listed  in  Table  1.  The  MDL  for a specific  wastewater may differ
from  that  listed, depending  upon  the nature  of interferences  in  the sample
matrix.

      1.4   This  method  is  restricted to-  use  by or under the  supervision of
analysts experienced in the use  of gas chromatography and in the interpretation
of gas chromatograms.   Each analyst must  demonstrate  the  ability  to  generate
acceptable results with this method using the procedure described in  Section 8.2.

      1.5   The toxicity or carcinogenicity of each  reagent used  in this method
has  not  been precisely  defined.   However,  each chemical  compound should be
treated as a potential  health hazard.   From this viewpoint,  exposure  to these
chemicals  must be  reduced to  the lowest  possible level  by whatever  means
available.  The  laboratory  is  responsible for maintaining  a current awareness
file of OSHA regulations regarding  the safe handling of the chemicals specified
in this method.  A reference file of material data  handling sheets should also

                                   8110 -  1                         Revision 0
                                                                     July 1992

-------
be made available to all personnel involved in the chemical analysis.  Additional
references to laboratory safety are available and  have been identified.


2.0   SUMMARY OF METHOD

      2.1   A measured  volume  of  sample,  approximately one-liter,  is  solvent
extracted with  methylene  chloride  using a  separatory funnel.   The methylene
chloride extract  is  dried and exchanged to hexane during concentration  to  a
volume of 10 mL or less. GC conditions are described which permit the separation
and measurement  of the compounds in  the extract using a  halide specific detector.

      2.2   Method 8110 provides gas chromatographic conditions for the detection
of ppb concentrations of haloethers.   Prior to use of this method,  appropriate
sample extraction techniques must be used.  Both neat and diluted organic liquids
(Method 3580, Waste  Dilution)  may be  analyzed by direct injection.  A 2 to 5 jiL
aliquot  of  the  extract is  injected  into a  gas  chromatograph (GC)  using the
solvent  flush technique,  and  compounds  in  the GC effluent are detected  by an
electrolytic conductivity detector (HECD).


3.0    INTERFERENCES

      3.1   Refer to Methods 3500,  3600, and 8000.

      3.2   Matrix  interferences   may   be   caused  by  contaminants  that  are
coextracted  from  the sample.   The extent  of matrix interferences  will  vary
considerably from source to source, depending upon the nature and diversity of
the industrial  complex or municipality being sampled.  The cleanup procedures in
Section  7.3  can be  used to  overcome  many  of these  interferences,  but  unique
samples may require additional cleanup approaches to achieve the MDL listed in
Table  1.

       3.3   Dichlorobenzenes are known to coelute with haloethers  under some gas
chromatographic conditions.  If these materials  are present in a sample, it may
be  necessary to  analyze  the  extract  with  two  different  column  packings to
completely  resolve all  of the compounds.

       3.4   Solvents, reagents, glassware,  and other sample processing hardware
may yield discrete artifacts and/or elevated baselines  causing misinterpretation
of gas chromatograms.   All these materials must be demonstrated to  be free from
interferences under the conditions  of the analysis, by  analyzing reagent blanks.
Specific selection of reagents and purification of solvents by distillation in
all-glass systems may be required.


4.0    APPARATUS AND MATERIALS

       4.1   Gas chromatograph

            4.1.1 Gas   chromatograph  -  An  analytical  system  complete  with
       temperature  programmable  gas   chromatograph  suitable  for  on-column
       injection  and  all  required  accessories  including  syringes, analytical
       columns,  gases,  detector,  and strip-chart  recorder.  A data system is

                                   8110 - 2                          Revision  0
                                                                     July  1992

-------
      recommended for measuring peak areas,

            4.1.2 Columns

                  4.1.2.1     Column 1  -  1.8  m x 2 mm  ID  pyrex  glass,  packed
            with  Supelcoport,  (100/120  mesh)  coated  with  3%  SP-1000  or
            equivalent.  This column was used  to develop the method performance
            statements in  Section  9.0.    Guidelines  for the use  of  alternate
            column packings are provided in Section 7.3.1.

                  4.1.2.2     Column 2  -  1.8  m x 2 mm  ID  pyrex  glass,  packed
            with  2,6-diphenylene  oxide  polymer  {Tenax-GC  60/80  mesh)  or
            equivalent.

            4.1.3 Detector  - Electrolytic  conductivity or  microcoulometric.
      These detectors have proven effective in the analysis of wastewaters for
      the parameters listed in the  scope of this method.  The Hall conductivity
      detector (HECD) was used to develop the  method performance statements in
      Section 9.0.  Guidelines for the use of  alternate detectors are provided
      in Section 7.3.1.  Although less selective, an electron capture detector
      (ECD) is an acceptable alternative.

      4.2   Kuderna-Danish (K-D) apparatus

            4.2.1 Concentrator tube -  10 ml graduated (Kontes K-570050-1025 or
      equivalent).  A  ground glass stopper is used to  prevent  evaporation of
      extracts.

            4.2.2 Evaporation  flask  -   500   mL   (Kontes   K-570001-0500  or
      equivalent).   Attach  to  concentrator  tube  with springs,  clamps,  or
      equivalent.

            4.2.3 Snyder column  -  Three  ball  macro (Kontes  K-503000-0121  or
      equivalent).

            4.2.4 Springs -  1/2 inch  (Kontes  K-662750 or equivalent).

      4.3   Vials - Amber glass,  10 to 15  ml capacity, with Teflon lined screw-
cap or crimp top.

      4.4   Boiling  chips  -  Approximately 10/40  mesh.   Heat  to  400°C  for
30 minutes or Soxhlet extract with methylene chloride.

      4.5   Water  bath  -  Heated,  with 'concentric  ring  cover,  capable  of
temperature control (± 2°C).   The bath should  be  used  in a  hood.

      4.6   Balance - Analytical, 0.0001 g.

      4.7   Volumetric flasks, Class  A  - Appropriate sizes  with  ground glass
stoppers.
                                   8110 - 3                         Revision 0
                                                                     July 1992

-------
5.0   REAGENTS

      5.1   Reagent grade inorganic chemicals shall  be used in all tests.  Unless
otherwise indicated, it  is intended that all  inorganic reagents shall conform to
the specifications  of the  Committee  on Analytical Reagents of the  American
Chemical Society,  where  such specifications  are available.  Other grades may be
used,  provided it  is first ascertained  that  the reagent is of sufficiently high
purity to permit its use without lessening the accuracy of the determination.

      5.2   Organic-free reagent water  -  All  references to water in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.3   Acetone, CH3COCH3  -  Pesticide quality or equivalent.

      5.4   Hexane, C6HU - Pesticide quality or equivalent.

      5.5   Isooctane, (CH3)3CCH2CH(CH3)2 - Pesticide quality  or equivalent.

      5.6   Stock standard solutions (1000 mg/L) -  Stock standard solutions can
be prepared from pure standard materials or purchased  as  certified solutions.

            5.6.1  Prepare  stock  standard   solutions  by  accurately  weighing
      0.1000 ± 0.0010 g  of  pure  material.   Dissolve the  material  in pesticide
      quality acetone and dilute to volume in a 100  ml volumetric flask.  Larger
      volumes can  be used at the convenience of the  analyst.   If compound purity
      is certified at 96% or greater, the weight can be used without correction
      to  calculate the  concentration  of  the stock  standard.    Commercially
      prepared  stock  standards  can be  used  at  any concentration  if  they are
      certified by the manufacturer or by an independent  source.

            5.6.2 Transfer the stock standard solutions into bottles with Teflon
      lined screw-caps  or crimp tops.    Store at   4°C  and  protect  from light.
      Stock  standard solutions  should  be   checked  frequently  for  signs  of
      degradation or evaporation, especially just prior to preparing calibration
      standards from them.

            5.6.3 Stock standard solutions must be replaced after six months, or
      sooner if comparison with check  standards indicates a problem.

      5.7   Calibration  standards - Calibration standards at a minimum of five
concentrations should be prepared through dilution of the stock standards with
isooctane.  One of the  concentrations  should be at a  concentration near, but
above,  the  method  detection  limit.    The  remaining  concentrations  should
correspond  to  the expected range  of concentrations  found in real  samples or
should  define  the  working  range of the GC.   Calibration  solutions  must be
replaced after six months, or sooner if  comparison with  check standards  indicates
a problem.

      5.8   Internal standards  (if  internal  standard calibration is used) - To
use'this approach,  the analyst must select  one or  more internal standards that
are similar in  analytical behavior to" the compounds  of interest.   The analyst
must further demonstrate that the  measurement of  the  internal standard is not
affected by method  or matrix  interferences.  Because  of these limitations, no
internal standard can be  suggested  that  is  applicable to all  samples.

                                   8110  - 4                         Revision 0
                                                                      July  1992

-------
            5.8.1 Prepare  calibration   standards  at   a  minimum  of   five
      concentrations for each analyte of interest as described in Section 5.7.

            5,8.2 To each calibration standard, add a known constant amount of
      one or more internal standards, and dilute to volume with isooctane.

            5.8.3 Analyze each calibration standard according to Section  7.0.

      5.9   Surrogate standards -  The analyst  should monitor the performance of
the extraction,  cleanup  (when used), and analytical system and the effectiveness
of  the  method  in  dealing with  each  sample  matrix by  spiking each  sample,
standard, and reagent blank with one or two surrogates {e.g. haloethers that are
not expected  to  be  in the sample)  recommended to encompass the range of the
temperature program  used in this method.  Method 3500 details instructions on the
preparation of base/neutral  surrogates.  Deuterated analogs of analytes should
not be  used  as  surrogates for  gas  chromatographic analysis due  to coelution
problems.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See the  introductory  material  to this chapter,  Organic Analytes,
Section 4.1.   Extracts  must  be stored  at  4°C and analyzed within  40  days of
extraction.


7.0   PROCEDURE

      7.1   Extraction

            7.1.1 Refer to Chapter Two for guidance on choosing the appropriate
      extraction procedure.    In  general, water  samples  are  extracted   at  a
      neutral, or as is, pH with methylene chloride,  using either Method 3510 or
      3520.  Solid samples are extracted using either Method 3540 or 3550.

            NOTE: Some  of the  haloethers  are  very  volatile  and  significant
                  losses  will  occur  in concentration  steps  if  care  is  not
                  exercised.    It  is  important  to maintain a  constant  gentle
                  evaporation rate and  not to allow the  liquid volume  to fall
                  below 1 to  2 mL before removing the K-D  apparatus from the hot
                  water bath.

            7.1.2 Prior to gas  chromatographic analysis,  the extraction solvent
      must be exchanged to hexane.   The exchange is performed  during  the K-D
      procedures listed  in  all of  the extraction methods.   The exchange is
      performed as follows.

                  7.1.2.1     Following K-D of the methylene chloride extract to
            1 mL using the macro-Snyder column,  allow the apparatus  to cool and
            drain for at least 10 minutes.

                  7.1.2.2     Momentarily remove the Snyder column,  add  50 mL of
            hexane,  a new boiling  chip,  and  reattach  the macro-Snyder  column.
            Concentrate the extract  using  1 mL of hexane to prewet  the Snyder

                                   8110 - 5                         Revision 0
                                                                     July  1992

-------
      column.   Place the K-D  apparatus  on the  water  bath so  that  the
      concentrator tube  is partially  immersed  in the hot  water.  Adjust
      the vertical position of the apparatus and the water temperature, as
      required, to complete  concentration in 5-10 minutes.  At the proper
      rate of distillation the  balls of the column will  actively chatter,
      but the chambers will not flood.  When the apparent volume  of liquid
      reaches  1 ml,  remove  the K-D apparatus  and allow  it  to drain and
      cool  for at  least  10  minutes.    The  extract  will   be  handled
      differently  at this point, depending on  whether  or  not  cleanup is
      needed.  If  cleanup is not required, proceed to Section 7.1.2.3.  If
      cleanup  is needed, proceed to Section 7.1.2.4.

            7.1.2.3      If cleanup of the extract is not required, remove
      the Snyder column  and rinse the flask and  its lower joint into the
      concentrator  tube with  1-2 ml  of  hexane.   A  5  ml  syringe  is
      recommended  for this  operation.    Adjust  the  extract  volume  to
      10.0 ml.  Stopper  the  concentrator tube  and store  refrigerated at
      4°C if further  processing will not be performed immediately.  If the
      extract  will  be   stored  longer   than   two  days,  it  should  be
      transferred  to  a  Teflon  lined  screw-cap vial.   Proceed  with  gas
      chromatographic analysis.

            7.1.2.4      If cleanup of the extract is required,  remove the
      Snyder  column  and rinse the  flask and  its lower  joint  into  the
      concentrator tube with a minimum amount of hexane.   A 5  ml syringe
      is recommended for this operation.   Add a clean boiling chip to the
      concentrator tube  and attach a two ball micro-Snyder column.  Prewet
      the column by adding about 0.5 ml of hexane to  the top.   Place the
      micro-K-D  apparatus  on   the   water  bath  (80°C)   so   that  the
      concentrator tube  is partially  immersed  in the hot  water.  Adjust
      the vertical position of the apparatus and the water temperature, as
      required, to complete  concentration in 5-10 minutes.  At the proper
      rate of distillation the  balls of the column will  actively chatter,
      but the chambers will not flood.  When the apparent volume of liquid
      reaches  0.5 ml, remove the K-D apparatus and allow it to drain and
      cool for at least  10 minutes.

            7.1.2.5      Remove the micro-Snyder column and rinse  the flask
      and  its lower  joint  into the  concentrator tube with  0.2  ml  of
      hexane.  Adjust the extract volume to 2.0 ml and proceed with either
      Method  3610 or 3620.

7.2   Cleanup

      7.2.1 Proceed  with  Method 3620,  using  the 2  ml  hexane  extracts
obtained from  Section 7.1.2.5.

      7.2.2 Following cleanup,  the extracts should be analyzed by GC, as
described in  the previous paragraphs and in Method 8000.

7.3   Gas Chromatography Conditions

      7.3.1 Table  1  summarizes  the  recommended operating  conditions for
the gas chromatograph.  This  table  includes retention times and MDLs that

                             8110 - 6                         Revision 0
                                                               July 1992

-------
      were  obtained  under  these  conditions.  ,  Examples  of  the  parameter
      separations achieved by these  columns are shown in Figures 1  and 2.  Other
      packed columns, chromatographic conditions, or  detectors  may be  used if
      the requirements of Section 8.2 are met.  Capillary (open-tubular) columns
      may also  be used  if the  relative standard deviations  of responses for
      replicate  injections  are  demonstrated  to  be  less  than   6%  and  the
      requirements of Section 8.2 are met.

      7.4   Calibration - Refer to Method 8000 for proper calibration techniques.
Use Table 1  and  especially Table 2 for guidance on  selecting the  lowest point on
the calibration curve.

            7.4.1 The procedure  for internal  or  external  calibration  may be
      used.   Refer to Method 8000 for a description of  each of these procedures.

            7.4.2 If  cleanup  is  performed on  the samples, the  analyst  should
      process a series  of  standards  through  the  cleanup  procedure and  then
      analyze the  samples by  GC.   This will confirm  elution  patterns  and the
      absence of interferents from the reagents.

      7.5   Gas chromatographic analysis

            7.5.1 Refer  to Method 8000.   If the  internal  standard calibration
      technique is used,  add 10 nL of internal  standard to the sample prior to
      injection.

            7.5.2 Method  8000 provides  instructions on  the  analysis sequence,
      appropriate  dilutions,  establishing daily  retention  time  windows,  and
      identification criteria.   Include a mid-concentration check standard after
      each group of  10 samples in the analysis sequence.

            7.5.3 Examples of GC/HECD  chromatograms for haloethers are shown in
      Figures 1 and 2.

            7.5.4 Record the  sample  volume injected and the resulting peak sizes
      (in area units or peak heights).

            7.5.5 Using  either  the  internal or external  calibration procedure
      (Method 8000),  determine the identity and quantity of each  analyte peak in
      the sample chromatogram.  See Method 8000  for calculation equations.

            7.5.6 If  peak detection and  identification  are  prevented  due to
      interferences,  the hexane  extract  may undergo cleanup using either Method
      3610 or 3620.


8.0   QUALITY CONTROL

      8.1   Refer  to Chapter  One  for  specific  quality  control  procedures.
Quality control  to validate sample extraction  is  covered in Method 3500 and in
the extraction method utilized.  If extract cleanup was performed,  follow the QC
in Method 3600 and in the specific cleanup method.
                                   8110 - 7                         Revision 0
                                                                     July 1992

-------
      8.2   Procedures to check the GC system operation are found in Method 8000,
Section 8.6.

            8.2.1 The quality control  (QC)  reference sample concentrate (Method
      8000, Section 8.6) should contain each analyte of interest at 20 mg/L.

            8.2.2 Table  1   indicates  the   recommended  operating  conditions,
      retention  times,  and  MDLs  that were  obtained  under  these  conditions.
      Table 2 gives method  accuracy and precision  for the analytes of interest.
      The contents  of both Tables  should  be  used to evaluate a laboratory's
      ability to perform and generate acceptable data by this method.

      8.3   Calculate surrogate standard recovery on  all  samples,  blanks,  and
spikes.   Determine if  the  recovery  is within  limits  (limits  established  by
performing QC procedures outlined in Method 8000, Section 8.10).

            8.3.1 If recovery is not within limits, the following is required.

            •     Check  to  be sure that there are no  errors in calculations,
                  surrogate  solutions and  internal  standards.   Also,  check
                  instrument performance.

            •     Recalculate the data and/or reanalyze the  extract if any of
                  the above checks reveal  a problem.

            •     Reextract and reanalyze  the sample  if  none  of the above are a
                  problem or  flag the data as "estimated concentration."


9.0   METHOD PERFORMANCE

      9.1   This method  has  been  tested for linearity of recovery from spiked
organic-free reagent  water  and  has  been demonstrated  to be applicable for the
concentration range from 4 x MOL to 1000 x MDL.

      9.2   In  a single laboratory  (Monsanto  Research  Center),  using spiked
wastewater samples, the  average recoveries presented in Table  2 were obtained.
Each spiked sample was analyzed in triplicate on  three  separate occasions. The
standard deviation of the percent recovery  is also included  in  Table 2.


10.0  REFERENCES

1.    Fed. Reoist. 1984, 49_,  43234; October 26.

2.    Mills, P.A. "Variation  of Florisil Activity: Simple Method for Measuring
      Absorbent Capacity and Its Use in Standardizing Florisil Columns"; Journal
      of the Association of  Official  Analytical Chemists 1968,  51, 29.

3.    Handbook   of   Analytical   Quality   Control   in   Water   and  Wastewater
      Laboratories; U.S. Environmental Protection  Agency. Office of Research and
      Development.   Environmental   Monitoring   and  Support   Laboratory.  ORD
      Publication  Offices  of  Center for  Environmental  Research Information:
      Cincinnati, OH, 1979;  EPA-600/4-79-019.

                                   8110 -  8                         Revision 0
                                                                     July 1992

-------
4.    Methods for  Chemical  Analysis of  Water and  Wastes;  U.S.  Environmental
      Protection Agency.  Office  of Research  and  Development.  Environmental
      Monitoring and Support Laboratory. ORD Publication Offices of Center for
      Environmental Research Information:  Cincinnati, OH, 1983; EPA-600/4-79-
      020.

5.    Burke,  J.A.  "Gas  Cnromatography for  Pesticide Residue  Analysis;  Some
      Practical  Aspects";  Journal of  the  Association o_f  Official  Analytical
      Chemists 1965, 48, 1037.

6.    "EPA Method Validation Study 21 Methods 611  (Haloethers)," Report for EPA
      Contract 68-03-2633.

7.    "Determination of  Haloethers in Industrial  and Municipal Wastewaters";
      Report for EPA Contract 68-03-2633 (In preparation).
                                    8110  -  9                          Revision  0
                                                                      July 1992

-------
                                   TABLE 1.
            CHROMATOGRAPHIC CONDITIONS AND METHOD DETECTION LIMITS
Analyte
       Retention Time
          (minutes)	
     Column 1     Column 2
    Method
Detection Limit
    (M9/L)
Bis(2-chloroisopropyl) ether
Bis(2-chloroethyl) ether
Bis(2-chloroethoxy)methane
4-Chlorophenyl phenyl ether
4-Bromophenyl phenyl ether
8.4
9.4
13.1
19.4
21.2
9.7
9.1
10.0
15.0
16.2
0.8
0.3
0.5
3.9
2.3
Column 1 conditions:
   Carrier gas (He) flow rate:
   Initial temperature:
   Temperature program:
   Final temperature:
       40 mL/min
       60°C, hold for 2 minutes
       60°C to 230°C at 8°C/min
       230°C, hold for 4 minutes
   Under these conditions the retention time for aldrin is 22.6 minutes.
Column 2 conditions:
   Carrier gas (He) flow
   Initial temperature:
   Temperature program:
   Final temperature:
rate:  40 mL/min
       150°C, hold for 4 minutes
       150°C to 310°C at 16°C/nnn
       310°C
   Under these conditions the retention time for aldrin  is  18.4 minutes.
                                   8110 - 10
                                           Revision  0
                                            July  1992

-------
                                   TABLE 2.
                    SINGLE OPERATOR ACCURACY AND PRECISION
                             Average   Standard    Spike    Number
                             Percent   Deviation   Range      of     Matrix
Analyte                      Recovery      %       (M9/L)  Analyses  Types

Bis(2-chloroethoxy)methane62571f38273
Bis(2-chloroethyl) ether         59       4.5         97        27       3
Bis(2-chloroisopropyl) ether     67       4.0         54        27       3
4-Bromophenyl phenyl ether       78       3.5         14        27       3
4-Chlorophenyl phenyl ether      73       4.5         30        27       3
                                   8110 - 11                         Revision  0
                                                                      July  1992

-------
              FIGURE  1.
  GAS CHROMATOGRAM OF HALOETHERS
Column:  3% SP-TOOO on Suptlcopon
Prognm: 60*C. -2 minutit i*/minut» to 23O*C.
Dtttctor: Hull tltctrolytic conductivity
 1C
   24   6  8   10  12  14  IS  18  20  22 24
                8110 -  12
Revision 0
 July 1992

-------
                FIGURE 2.
    GAS CHROMATOGRAM OF HALOETHERS
Column: T»n»x GC
Prognm: JSO°C.-4 minuttt 16"/minut» to 310°C.
Dtttftar: Hill ultctrolytie conductivity
                            I
9      12       16

 ftttintiofi timt.
                                  20
24
                8110  -  13
                                           Revision  0
                                            July 1992

-------
                         METHOD 8110
            HALOETHERS  BY  GAS CHROMATOGRAPHY
                               Start
  7124 Perform
nvicra-K-D procedure
   using hexan*.
proceed wilh Method
   3610 or  3620
Yes
                           7 I I ChooJQ
                            aopropria Ie
                            ett t rac 11 art
                             procadu r»
        712 Perform
      solvent exchange
        us ing hexane
                                  No
       7123 Adjust
     extract volume and
        pr oceed *itn
      ana Lysis or store
       in appropriate
          manner
                           7 3 1 Hcfer  '.:
                            Taole 1  :or
                            r ecommcnded
                             opera 11.19
                         conditions IDT  '.
                              7  4  Ref or  I o Me (.hod
                                30CO for proper
                                  c a I i-Dration
                                  techniques
                                                    7 S 1 Refer  to
                                                    Method 8000  for
                                                    guidance on  CC
                                                      anal/Si >
                              7 5  4  Record sample
                              volume injected and
                              resulting peak size
                                                     7 S S Perform
                                                     aopropridte
                                                    1cula tions  |ref (
                                                    to Method 30CO]
                                     5'. op
                            8110 -  14
                                                    Revision 0
                                                     July  1992

-------
00

-------
                                 METHOD 8120A

                CHLORINATED HYDROCARBONS BY GAS CHROMATOGRAPHY
1.0   SCOPE AND APPLICATION

      1.1   Method  8120 is  used to  determine  the
chlorinated hydrocarbons.   The following compounds
method:
                   concentration  of  certain
                  can  be  determined  by this
Compounds
          Appropriate Preparation Techniques

CAS No8     3510     3520  3540/   3550   3580
                          3541
2-Chloronaphthalene
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
Hexachlorobenzene
Hexachlorobutadiene
Hexachl orocycl ohexane
Hexachl orocycl opentadi ene
Hexachl oroethane
Pentachlorohexane
Tetrachl orobenzenes
1,2,4-Trichlorobenzene
91-58-7
95-50-1
541-73-1
106-46-7
118-74-1
87-68-3
608-73-1
77-47-4
67-72-1


120-82-1
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
ND
X
X
X
X
X
X
X
X
X
X
X
ND
X
X
X
X
X
X
X
X
X
X
X
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
   a  Chemical Abstract Services Registry Number.
   x  Greater than 70 percent recovery by this technique
   ND Not determined.

      1.2   Table 1 indicates compounds that  may be determined by this method and
lists the method  detection limit for each compound in organic-free reagent water.
Table 2 lists the estimated quantitation limit (EQL) for other matrices.


2.0   SUMMARY OF METHOD

      2.1   Method 8120 provides gas chromatographic conditions for the detection
of ppb concentrations of certain chlorinated hydrocarbons.  Prior to use of this
metho'd, appropriate  sample extraction  techniques must  be used.   Both  neat and
diluted organic liquids (Method 3580,  Waste  Dilution) may be analyzed by direct
injection.   A  2 to 5 juL aliquot  of the extract  is  injected  into a  gas
chromatograph (GC),  and compounds in the GC effluent are detected by an electron
capture detector (ECD).
                                  8120A  -  1
                                  Revision 1
                              September 1994

-------
      2.2   If interferences  are encountered in the  analysis,  Method 8120 may
also be performed on extracts that have undergone cleanup using Method 3620.

3.0   INTERFERENCES

      3.1   Refer to Methods  3500,  3600,  and 8000.

      3.2   Solvents,  reagents,  glassware, and other  sample processing hardware
may yield discrete artifacts and/or elevated  baselines causing misinterpretation
of gas chromatograms.  All of these  materials must  be  demonstrated to be free
from interferences, under the conditions  of the  analysis,  by analyzing method
blanks.    Specific selection  of  reagents   and  purification  of  solvents  by
distillation in all glass systems may be required.

      3.3   Interferences coextracted from samples  will  vary considerably from
source to  source,  depending  upon  the  waste being  sampled.   Although general
cleanup techniques  are recommended as  part  of this  method,  unique samples may
require additional cleanup.


4.0   APPARATUS AND MATERIALS

      4.1   Gas chromatograph

            4.1.1   Gas  chromatograph  - Analytical  system  complete   with  gas
      chromatograph  suitable  for   on-column  injections   and  all   required
      accessories, including detectors, column supplies,  recorder,  gases, and
      syringes.  A data system for measuring peak areas and/or peak heights is
      recommended.

            4.1.2  Columns

                   4.1.2.1     Column 1  -  1.8 m  x 2  mm  ID  glass  column packed
            with 1% SP-1000 on Supelcoport (100/120  mesh)  or equivalent.

                   4.1.2.2     Column 2  -  1.8 m  x 2  mm  ID  glass  column packed
            with  1.5%  OV-1/2.4%  OV-225  on  Supelcoport  (80/100  mesh)  or
            equivalent.
      4.2
4.1.3  Detector -  Electron  capture (ECD).

Kuderna-Danish (K-D) apparatus
            4.2.1   Concentrator tube - 10 ml,  graduated  (Kontes K-570050-1025 or
      equivalent).   A  ground  glass  stopper is used to  prevent  evaporation of
      extracts

            4.2.2   Evaporation   flask  -   500  ml  (Kontes   K-570001-500   or
      equivalent).    Attach  to  concentrator  tube with   springs,  clamps  or
      equivalent.

            4.2.3   Snyder  column  - Three  ball macro (Kontes  K-503000-0121 or
      equivalent).
                                   8120A  -  2
                                                        Revision 1
                                                    September 1994

-------
            4.2.4  Snyder  column  -  Two  ball  micro  (Kontes  K-569001-0219  or
      equivalent).

            4.2.5  Springs  -   1/2  inch  (Kontes K-662750  or  equivalent).

      4.3   Boiling chips - Solvent extracted,  approximately 10/40 mesh (silicon
carbide or equivalent).

      4.4   Water  bath -  Heated,  with  concentric  ring   cover,  capable  of
temperature control (+ 5°C).   The  bath  should  be used in a  hood.

      4.5   Volumetric flasks  -  10,  50,  and  100  ml, with ground glass stoppers.

      4.6   Microsyringe  -  10  juL.

      4.7   Syringe -  5 ml.

      4.8   Vials - Glass,  2,  10,  and  20  ml capacity with  Teflon  lined screw-
caps or crimp tops.


5.0   REAGENTS

      5.1   Reagent grade inorganic chemicals shall be used  in  all tests. Unless
otherwise  indicated,  it  is intended  that all  reagents  shall conform  to the
specifications of the  Committee  on Analytical Reagents of the American Chemical
Society, where such specifications  are available.   Other  grades  may be used,
provided it is first ascertained that the reagent is  of sufficiently high purity
to permit its use without lessening the accuracy of the determination.

      5.2   Organic-free  reagent water.  All references to water in this method
refer to organic-free reagent water, as defined  in Chapter One.

      5.3   Solvents

            5.3.1  Hexane, C6H14.   Pesticide  quality  or  equivalent.

            5.3.2  Acetone, CH3COCH3.  Pesticide  quality or equivalent.

            5.3.3  Isooctane,  C8H18.   Pesticide quality  or equivalent.

      5.4   Stock standard  solutions

            5.4.1  Prepare  stock standard  solutions  at  a concentration of 1000
      mg/l by dissolving  0.0100 g  of assayed reference material in  isooctane or
      hexane and  diluting to volume in a 10 ml volumetric flask.  Larger volumes
      can be used at the convenience of the analyst.   When compound  purity is
      assayed to  be 96% or greater, the weight  can  be  used without correction to
      calculate the concentration  of the stock standard.  Commercially prepared
      stock standards  can be used  at any concentration if they are  certified by
      the manufacturer or by  an  independent  source.
                                   8120A  - 3                         Revision 1
                                                                September 1994

-------
            5.4.2  Transfer  the stock standard solutions into vials with Teflon
      lined screw  caps  or crimp tops.   Store  at 4°C and  protect  from light.
      Stock standards should be checked frequently for signs of degradation or
      evaporation,  especially just prior to preparing calibration standards.

            5.4.3  Stock standard  solutions must be replaced after one year, or
      sooner if comparison with check standards  indicates  a problem.

      5.5   Calibration  standards  -  Calibration  standards  at a minimum of five
concentrations should be prepared through dilution of the stock standards with
isooctane or  hexane.  One of the concentrations  should be  at a concentration
near, but above,  the method detection limit.  The  remaining concentrations should
correspond  to  the  expected range of  concentrations  found  in  real  samples or
should define  the  working  range  of  the  GC.    Calibration solutions  must be
replaced after six months, or sooner  if comparison with check standards indicates
a problem.

      5.6   Internal  standards (if internal  standard  calibration is used) - To
use this approach,  the analyst must select one or more internal standards that
are similar in analytical behavior  to  the  compounds  of  interest.   The analyst
must further demonstrate  that the measurement of  the internal  standard is not
affected by method or matrix  interferences.   Because of these limitations, no
internal  standard can be suggested that is applicable to all samples.

            5.6.1  Prepare  calibration  standards   at   a  minimum   of   five
      concentrations for each analyte of interest as  described in Sec.  5.5.

            5.6.2  To each calibration  standard,  add a known constant amount of
      one or more  internal  standards,  and dilute to  volume with  isooctane or
      hexane.

            5.6.3  Analyze each calibration  standard  according  to Sec.  7.0.

      5.7   Surrogate  standards  -  The analyst should monitor the performance of
the extraction, cleanup  (when used), and analytical system and the effectiveness
of  the  method in  dealing  with  each  sample  matrix   by  spiking each  sample,
standard, and organic-free reagent water blank with one or two surrogates  (e.g.
chlorinated hydrocarbons that are  not  expected to  be  in the sample)  recommended
to encompass the range of the temperature program used in  this method.   Method
3500  details   instructions  on  the  preparation  of   base/neutral  surrogates.
Deuterated  analogs of  analytes  should not  be  used as  surrogates  for  gas
chromatographic  analysis due to coelution problems.


6.0   SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1   See the introductory material to this  chapter, Organic Analytes, Sec.
4.1.

      6.2   Extracts  must be stored  under refrigeration  and analyzed  within 40
days of extraction.
                                   8120A  -  4                         Revision 1
                                                                September 1994

-------
7.0   PROCEDURE

      7.1   Extraction

            7.1.1  Refer to Chapter Two for guidance  on choosing the appropriate
      extraction procedure.    In  general, water  samples  are  extracted  at  a
      neutral,  or as is, pH with methylene chloride,  using either Method 3510 or
      3520.   Solid samples are extracted using either Methods 3540/3541 or 3550.

            7.1.2  Prior to gas chromatographic analysis,  the extraction solvent
      must be exchanged to  hexane.   The exchange is performed  during  the K-D
      procedures listed in  all of  the extraction  methods.   The  exchange  is
      performed as  follows.

                  7.1.2.1     Following K-D  of  the  methylene  chloride  extract
            to  1 ml  using  the macro  Snyder column,  allow the  apparatus  to cool
            and  drain for  at least  10  minutes.

                  7.1.2.2     Momentarily remove the Snyder column,  add 50 ml
            of  hexane, a new boiling chip, and reattach the macro Snyder column.
            Concentrate  the  extract  using 1  ml  of hexane to prewet the Snyder
            column.   Place  the K-D apparatus  on  the water  bath   so that the
            concentrator tube is partially immersed  in the hot  water.   Adjust
            the  vertical position of the apparatus  and the water temperature, as
            required, to complete concentration  in 5-10 minutes.  At the proper
            rate of  distillation the balls of the column will  actively chatter,
            but  the chambers will not flood.   When  the apparent volume of liquid
            reaches  1  ml,  remove the  K-D apparatus  and   allow  it  to  drain and
            cool  for  at  least  10  minutes.    The   extract  will  be  handled
            differently  at this point, depending on whether or  not cleanup is
            needed.   If cleanup is not required, proceed to Sec.  7.1.2.3.  If
            cleanup  is  needed,  proceed to Sec.  7.1.2.4.

                  7.1.2.3     If cleanup of the  extract is not required, remove
            the  Snyder  column and  rinse the  flask  and its lower joint into the
            concentrator  tube  with  1-2  ml  of  hexane.    A  5  ml syringe  is
            recommended  for  this  operation.    Adjust  the extract  volume  to
            10.0 ml.  Stopper the concentrator tube and store refrigerated at 4°C
            if  further  processing will  not  be performed immediately.   If the
            extract  will  be   stored   longer than  two  days,  it  should  be
            transferred  to a vial  with a Teflon lined screw cap or crimp top.
            Proceed  with gas chromatographic  analysis.

                  7.1.2.4     If cleanup of the  extract is required, remove the
            Snyder  column  and  rinse the  flask  and   its  lower joint into the
            concentrator tube with  a minimum amount  of hexane.   A  5 ml  syringe
            is  recommended for  this  operation.  Add  a clean boiling chip to the
            concentrator tube and attach a two ball micro  Snyder column.   Prewet
            the  column  by  adding  about 0.5 ml of hexane  to the top.  Place the
            micro K-D apparatus on the  water bath (80°C) so that the concentrator
            tube is  partially immersed in the hot water.   Adjust  the vertical
            position of the apparatus and the water temperature, as required, to
            complete  concentration  in  5-10  minutes.   At  the  proper  rate  of


                                  8120A  - 5                         Revision 1
                                                                September 1994

-------
            distillation  the  balls of the column will actively chatter, but the
            chambers will not flood.  When the  apparent volume of liquid reaches
            0.5  mL, remove  the K-D apparatus and allow it to drain and cool for
            at least  10 minutes.

                  7.1.2.5     Remove the micro Snyder column and rinse the flask
            and  its  lower  joint  into  the concentrator  tube with  0.2 ml  of
            hexane.  Adjust the extract volume  to 2.0 ml and proceed with Method
            3620.

      7.2   Gas  chromatographic  conditions  (Recommended)

            7.2.1  Column 1

            Carrier gas  (5% methane/95% argon)  flow  rate  =  25 mL/min
            Column temperature =    65°C isothermal, unless otherwise specified
                                    (see Table 1).

            7.2.2  Column 2

            Carrier gas  (5% methane/95% argon)  flow  rate  =  25 mL/min
            Column temperature =    75°C isothermal, unless otherwise specified
                                    (see Table 1).

      7.3   Calibration - Refer to Method 8000 for proper calibration techniques.
Use Table 1  and  especially Table 2 for guidance on selecting the lowest point on
the calibration  curve.

            7.3.1  The procedure for internal  or  external  calibration  may  be
      used.   Refer to  Method 8000 for a  description of each of these procedures.

            7.3.2  If  cleanup is performed on the samples, the  analyst should
      process a  series  of standards  through  the cleanup  procedure  and  then
      analyze the samples by GC.  This  will validate elution patterns and the
      absence of interferents from the  reagents.

      7.4   Gas  chromatographic  analysis

            7.4.1  Refer  to Method  8000.  If  the internal standard calibration
      technique  is  used,  add  10  /uL of internal standard to the  sample prior to
      injecting.

            7.4.2  Method 8000 provides  instructions on  the analysis  sequence,
      appropriate dilutions,  establishing  daily retention  time  windows,  and
      identification  criteria.   Include  a mid-concentration standard after each
      group  of 10 samples in  the analysis sequence.

            7.4.3  Examples  of  GC/ECD   chromatograms for certain  chlorinated
      hydrocarbons  are shown  in  Figures 1 and  2.

            7.4.4  Record the sample volume  injected and  the  resulting peak sizes
      (in area units  or peak  heights).
                                   8120A  - 6                         Revision 1
                                                                September 1994

-------
            7.4.5  Using  either  the  internal  or external  calibration procedure
      (Method 8000),  determine the identity and  quantity of each component peak
      in the sample  chromatogram which  corresponds to the  compounds  used for
      calibration  purposes.   See Method  8000  for calculation equations.

            7.4.6  If peak detection and  identification  are prevented  due  to
      interferences,  the  hexane  extract  may undergo cleanup using Method 3620.

      7.5   Cleanup:  If required, the samples  may be cleaned up using the Methods
presented in Chapter 4.

            7.5.1  Proceed  with  Method  3620   using  the 2  mL  hexane  extracts
      obtained from  Sec.  7.1.2.5.

            7.5.2  Following  cleanup, the extracts should be analyzed by GC,  as
      described in the previous  paragraphs  and in Method 8000.


8.0   QUALITY CONTROL

      8.1   Refer  to  Chapter One   for  specific  quality  control  procedures.
Quality control to validate sample  extraction is covered  in Method 3500 and  in
the extraction method utilized.  If extract cleanup was  performed, follow the QC
in Method 3600 and in the specific  cleanup method.

      8.2   Procedures to check the GC system operation are  found in Method 8000.

            8.2.1  The quality  control  check  sample concentrate  (Method 8000)
      should contain  each  parameter  of interest  at the following concentrations
      in acetone:  hexachloro-substituted hydrocarbon,  10  mg/L; and  any other
      chlorinated  hydrocarbon,  100  mg/L.

            8.2.2  Table 3 indicates  the calibration and QC acceptance criteria
      for this  method.  Table 4 gives method accuracy and precision as functions
      of concentration for  the  analytes  of   interest.   The contents of  both
      Tables should  be used to  evaluate a  laboratory's ability to  perform and
      generate acceptable data by this  method.

      8.3   Calculate surrogate  standard recovery on all  samples,  blanks,  and
spikes.   Determine  if the  recovery is within  limits  (limits  established  by
performing QC procedures  outlined in Method 8000).

            8.3.1  If recovery is not within limits, the  following procedures are
      required.

                  •     Check to be  sure there  are no  errors in calculations,
                        surrogate  solutions   and  internal  standards.    Also,
                        check instrument performance.

                  •     Recalculate  the data and/or  reanalyze  the  extract  if
                        any of  the above checks reveal  a problem.
                                  8120A  -  7                         Revision 1
                                                                September 1994

-------
                        Reextract and reanalyze the sample if none of the above
                        are   a   problem  or  flag  the   data  as   "estimated
                        concentration".
9.0   METHOD PERFORMANCE

      9.1   The method  was  tested by 20 laboratories using organic-free reagent
water, drinking water,  surface water, and three industrial wastewaters spiked at
six concentrations over the range 1.0 to 356 p,g/l.  Single operator precision,
overall precision, and  method accuracy were  found  to be directly related to the
concentration of the  parameter and essentially independent of  the sample matrix.
Linear equations to describe these relationships for a  flame ionization detector
are presented in Table 4.

      9.2   The accuracy  and precision obtained will be determined by the sample
matrix, sample preparation technique, and calibration procedures used.


10.0  REFERENCES

1,    "Development and  Application of Test Procedures for  Specific Organic Toxic
      Substances in  Wastewaters.   Category  3  - Chlorinated  Hydrocarbons,  and
      Category 8 - Phenols,"  Report for EPA  Contract 68-03-2625.

2.    Burke,  J.A.  "Gas  Chromatography  for   Pesticide Residue Analysis;  Some
      Practical Aspects,"  Journal  of  the Association of  Official  Analytical
      Chemists, 48,  1037, 1965.

3.    "EPA Method Validation  Study 22, Method 612   (Chlorinated Hydrocarbons),"
      Report for EPA Contract 68-03-2625.

4.    "Method  Performance  for  Hexachlorocyclopentadiene   by  Method  612,"
      Memorandum  from   R.  Slater,   U.S.   Environmental   Protection  Agency,
      Environmental  Monitoring and Support Laboratory, Cincinnati,  Ohio 45268,
      December 7,  1983.

5.    U.S. EPA 40 CFR Part 136, "Guidelines Establishing Test  Procedures for the
      Analysis of Pollutants  Under the Clean Water Act;  Final Rule and  Interim
      Final Rule and Proposed Rule," October 26,  1984.

6.    "Determination of  Chlorinated  Hydrocarbons   in  Industrial  and  Municipal
      Wastewaters,"  Report for EPA Contract  68-03-2625.
                                   8120A  - 8                         Revision 1
                                                                September 1994

-------
                                   TABLE 1.
                GAS CHROMATOGRAPHY OF CHLORINATED HYDROCARBONS
Compound
Retention time (min)

Col. 1       Col. 2
  Method
 Detection
limit (M9/I-)
2-Chloronaphthalene
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
Hexachlorobenzene
Hexachl orobutadi ene
Hexachl orocycl ohexane
Hexachl orocycl opentadi ene
Hexachl oroethane
Pentachlorohexane
Tetrachlorobenzenes
1,2,4-Trichlorobenzene
2.7a
6.6
4.5
5.2
5.6a
7.7

ND
4.9

--
15.5
3.6b
9.3
6.8
7.6
10. lb
20.0

16. 5C
8.3


22.3
0.94
1.14
1.19
1.34
0.05
0.34

0.40
0.03
--
--
0.05
   ND = Not determined.

   8150°C  column  temperature.

   b!65°C  column  temperature.

   C100°C  column  temperature.
                                   8120A - 9
                                    Revision 1
                                September 1994

-------
                                TABLE 2.
                 DETERMINATION OF  ESTIMATED  QUANTITATION
                   LIMITS (EQL) FOR VARIOUS MATRICES"
Matrix                                                             Factor
Ground water                                                            10
Low-concentration soil by ultrasonic extraction with GPC cleanup       670
High-concentration soil and sludges by ultrasonic extraction        10,000
Non-water miscible waste                                           100,000
   a  EQL = [Method detection limit  (see  Table  1)]  X [Factor found in this
      table].   For non-aqueous  samples,  the  factor is on a wet weight basis.
      Sample EQLs  are  highly matrix  dependent.   The EQLs  to be determined
      herein are provided for guidance and may not always be achievable.
                               8120A - 10                        Revision 1
                                                             September 1994

-------
                       TABLE 3.
                 QC  ACCEPTANCE  CRITERIA"


Parameter
2-Chloronaphthalene
1,2-Dichlorobenzene
1 , 3-Di chl orobenzene
1,4-Dichlorobenzene
Hexachl orobenzene
Hexachlorobutadiene
Hexachl orocycl opentadi ene
Hexachl oroethane
1, 2, 4-Tri chl orobenzene
Test
cone.
(M9/L)
100
100
100
100
10
10
10
10
100
Limit Range
for s for x
(M9/L) (M9/L)
37.3 29.5-126.9
28.3 23.5-145.1
26.4 7.2-138.6
20.8 22.7-126.9
2.4 2.6-14.8
2.2 D-12.7
2.5 D-10.4
3.3 2.4-12.3
31.6 20.2-133.7
Range
P. PS
(%)
9-148
9-160
D-150
13-137
15-159
D-139
0-111
8-139
5-149
s = Standard deviation of four recovery measurements, in /^g/L.
x = Average recovery
P,PS = Percent recovery
D = Detected; result
a Criteria from 40
for four recovery
measured.
measurements, in ng/L.



must be greater than zero.
CFR Part 136 for
Method 612. These cri
teria are
based directly upon the method performance data in Table 4.  Where
necessary, the  limits  for recovery have  been  broadened to assure
applicability of the limits  to  concentrations  below those used to
develop Table 4.
                      8120A  - 11                        Revision 1
                                                    September 1994

-------
                                   TABLE 4.
         METHOD ACCURACY AND  PRECISION AS  FUNCTIONS OF  CONCENTRATION8


Parameter
Chloronaphthalene
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
Hexachl orobenzene
Hexachlorobutadiene
Hexachl orocyclopentadiene"
Hexachl oroethane
1, 2, 4-Trichl orobenzene
Accuracy, as
recovery, x'
(M9/L)
0.75C+3.21
0.85C-0.70
0.72C+0.87
0.72C+2.80
0.87C-0.02
0.61C+0.03
0.47C
0.74C-0.02
0.76C+0.98
Single analyst
precision, sr'
(M9/L)
0.28X-1.17
0.22X-2.95
0.21X-1.03
0.16X-0.48
0.14X+0.07
O.lSx+0.08
0.24x
0.23X+0.07
0.23X-0.44
Overall
precision,
S' (/ig/L)
0.38X-1.39
0.41x-3.92
0.49X-3.98
0.35X-0.57
0.36X-0.19
0.53X-0.12
O.SOx
0.36X-0.00
0.40X-1.37
X'


V


S'


C

x
Expected  recovery  for  one  or  more  measurements  of  a  sample
containing a concentration of C, in /zg/L.

Expected  single  analyst standard deviation  of measurements at  an
average concentration of x, in /zg/L.

Expected  interlaboratory  standajrd  deviation  of measurements at  an
average concentration found of x, in /zg/L.

True value for the concentration, in /ug/L.

Average  recovery  found for  measurements of  samples  containing  a
concentration of C, in jug/L.

Estimates based upon the performance in a single laboratory.
                                  8120A - 12
                                                         Revision  1
                                                     September  1994

-------
                                   FIGURE  1
              Column:       1.5% OV-1  +1.5% OV-225 on Gas Chrom. Q
              Temperature:  75°C
              Detector:     Electron Capture
                                             I
                  t    «	t    i    «
                  4        I       12       16
                    RETENTION TIME (MINUTES)
20
Gas chromatagram of  chlorinated hydrocarbons  (high molecular weight compounds).
                                 8120A  -  13
               Revision 1
           September  1994

-------
                                     FIGURE 2
                      I I
                           Column:      1.5V OV-1 + 1.5% OV-225 on Gas Chrom. Q
                           Temperature: 160°C
                           Detector:    Electron Capture
                                             i    t    «
                   0        4        •       12      16
                          RETENTION TIME (MINUTES)
Gas chromatagram of chlorinated hydrocarbons (low molecular weight compounds).
                                     8120A  -  14
    Revision 1
September 1994

-------
                                   METHOD  8120A
              CHLORINATED  HYDROCARBONS BY  GAS  CHROMATOGRAPHY
  7.1.1 Choose
   appropriate
    extraction
  procedure (see
   Chapter 2).
 7.1.2 Exchange
extraction solvent
 to hexane during
 K-D procedures.
   7.2 Set gas
 chromatography
   conditions.
7.3 Refer to Method
  8000 for proper
    calibration
    techniques.
   7.3.2 Is
   cleanup
 necessary?
 7.3.2 Process a
series of standards
 through cleanup
procedure; analyze
     by GC.
7.4 Perform GC
 analysis (see
Method 8000).
                                        7.4.5
                                   Is identification
                                     & detection
                                     prevented by
                                    interferences?
                            7.5.1  Cleanup using
                               Method 3620.
                                    8120A  -  15
                                           Revision  1
                                      September  1994

-------
00

-------
                                  METHOD  8121

  CHLORINATED HYDROCARBONS BY GAS CHROMATOGRAPHY: CAPILLARY COLUMN TECHNIQUE


1.0   SCOPE AND APPLICATION

      1.1   Method 8121 describes the determination of chlorinated hydrocarbons
in extracts prepared from environmental samples and RCRA wastes.  It describes
wide-bore open-tubular, capillary column gas chromatography procedures using both
single column/single  detector  and dual-column/dual-detector  approaches.   The
following compounds can be determined by this method:


      Compound Name                           CAS Registry No.8
Benzal chloride
Benzotrichloride
Benzyl chloride
2-Chloronaphthalene
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
Hexachlorobenzene
Hexachl orobutadi ene
a-Hexachlorocyclohexane (a-BHC)
j6-Hexachlorocyclohexane (0-BHC)
7-Hexachlorocyclohexane (y-BHC)

-------
the nature of interferences in the sample matrix.  Table 2 lists the estimated
quantitation limits (EQL) for other matrices.

      1.4   Table 3 lists the compounds that have been determined by this method
and their retention times using the single column technique.  Table 4 lists dual
column/dual detector retention time  data.   Figures  1  and  2  are chromatograms
showing the single  column technique.  Figure 3 shows a chromatogram of the target
analytes eluted from a pair of DB-5/DB-1701 columns and detected with electron
capture detectors  (ECD) under the prescribed GC conditions listed in Table 2.

      1.5   This method  is  restricted to use  by or under the  supervision  of
analysts experienced in the use of a gas  chromatograph and in the interpretation
of gas chromatograms.


2.0   SUMMARY OF METHOD

      2.1   Method 8121 provides gas chromatographic conditions for the detection
of  ppb  concentrations  of chlorinated  hydrocarbons  in  water  and  soil  or ppm
concentrations in waste samples.  Prior to use of this method, appropriate sample
extraction techniques must be used for environmental samples (refer to Chapt. 2).
Both neat  and diluted  organic liquids  (Method  3580)  may be analyzed by direct
injection.  Spiked  samples are used  to  verify  the applicability of the chosen
extraction technique to each new sample type.  Analysis is accomplished by gas
chromatography utilizing an instrument equipped with wide bore capillary columns
and single or dual  electron capture detectors.


3.0   INTERFERENCES

      3.1   Refer to Methods 3500, 3600, and 8000.

      3.2   The  electron capture  detector  responds  to   all  electronegative
compounds. Therefore, interferences are possible by other halogenated compounds,
as  well  as phthalates and other  oxygenated compounds,  and,  organonitrogen,
organosulfur and organophosphorus compounds.  Second column confirmation or GC/MS
confirmation  are  necessary  to  ensure  proper  analyte identification  unless
previous characterization of the sample source will ensure proper identification.

      3.3   Contamination by carryover can occur whenever high-concentration and
low-concentration samples are sequentially analyzed.  To reduce carryover, the
syringe  used  for injection must be  rinsed  out between samples with  solvent.
Whenever  an extract  concentration  exceeds  that of  the  highest  calibration
standard,  it should be followed  by the analysis  of a  solvent blank to check for
cross-contamination.   Additional  solvent  blanks  interspersed with  the sample
extracts should be  considered whenever the analysis of a solvent blank indicates
cross-contamination problems.

      3.4   Phthalate  esters,  if present in a sample, will  interfere only with
the BHC  isomers because  they  elute  in Fraction  2 of  the  Florisil  procedure
described  in  Method 3620.  The  presence of phthalate esters can  usually  be
minimized  by  avoiding  contact  with  any  plastic  materials  and by  following
standard decontamination procedures of reagents and glassware.

                                   8121 - 2                         Revision 0
                                                                September 1994

-------
      3.5   The presence of elemental sulfur will result in large peaks,  and can
often mask the region of compounds eluting after 1,2,4,5-tetrachlorobenzene.  The
tetrabutylammonium  (TBA)-sulfite  procedure (Method  3660)  works well  for  the
removal of elemental sulfur.

      3.6   In  certain cases  some  compounds  coelute on  either one  or both
columns.   In  these cases  the  compounds must  be reported as  coeluting.   The
mixture can be reanalyzed by GC/MS techniques, see Sec. 8.7 and Method 8270.

            3.6.1 Using  the dual  column  system  of  analysis  the  following
      compounds coeluted:

            DB-5        1,4-dichlorobenzene/benzyl chloride
                        l,2,3,5-tetrachlorobenzene/l,2,4,5-tetrachlorobenzene
                        l,2,3,4-tetrachlorobenzene/2-chloronaphthalene

            DB-1701     benzyl  chloride/1,2-dichlorobenzene/hexachloroethane
                        benzal  chloride/1,2,4-trichlorobenzene/
                        hexachlorobutadiene

            Some of the injections showed a separation  of 1,2,4-trichlorobenzene
      from the other two compounds, however, this is not always  the case, so the
      compounds are listed as coeluting.

      3.7   Solvents,  reagents, glassware,  and other sample processing hardware
may yield discrete  artifacts and/or elevated baselines  causing misinterpretation
of gas chromatograms.   All  these materials  must be demonstrated to be free from
interferences under the conditions of the analysis,  by  analyzing reagent blanks.


4.0   APPARATUS AND MATERIALS

      4.1   Gas  chromatograph: An   analytical   system complete  with  a  gas
chromatograph suitable  for on-column  and  spl it-splitless injection,  and  all
required  accessories,  including syringes,  analytical   columns,  gases,  and  two
electron  capture detectors.  A  data  system for measuring peak  areas,  and dual
display of chromatograms is recommended.  A GC equipped with a  single GC column
and detector are acceptable, however,  second  column  confirmation is obviously
more time consuming.   Following are  the  single and  dual  column configurations
used for developing the retention time data  presented in the method.  The columns
listed  in the  dual column  configuration may  also  be used for single column
analysis.

            4.1.1 Single Column Analysis:

                  4.1.1.1     Column  1  -  30  m  x 0.53  mm   ID  fused-silica
            capillary  column  chemically  bonded  with trifluoropropyl  methyl
            silicone (DB-210 or equivalent).

                  4.1.1.2     Column  2  -  30  m  x 0.53  mm   ID  fused-silica
            capillary column chemically bonded with polyethylene glycol (DB-WAX
            or equivalent).


                                   8121 - 3                         Revision 0
                                                                September 1994

-------
            4.1.2 Dual Column Analysis:

                  4.1.2.1     Column  1  -   30  m  x  0.53  mm  ID  fused-silica
            open-tubular  column,  cross!inked  and  chemically bonded  with 95
            percent dimethyl and 5 percent  diphenyl-polysiloxane  (DB-5, RTx-5,
            SPB-5, or equivalent), 0.83 /xm  or  1.5 jum film thickness.

                  4.1.2.2     Column  2  -   30  m  x  0.53  mm  ID  fused-silica
            open-tubular  column  crosslinked   and  chemically  bonded  with 14
            percent  cyanopropylphenyl  and  86 percent  dimethyl-polysiloxane
            (DB-1701, RTX-1701,  or equivalent), 1.0  /xm film thickness.

            4.1.3 Splitter:  If  the splitter approach to dual column  injection
      is  chosen,  following  are  three  suggested  splitters.   An  equivalent
      splitter  is  acceptable.    See  Sec. 7.5.1  for a  caution  on  the  use of
      splitters.

                  4.1.3.1     Splitter  1  -  J&W Scientific press-fit Y-shaped
            glass 3-way union splitter  (J&W Scientific, Catalog no. 705-0733).

                  4.1.3.2     Splitter  2 -  Supelco  8 in.  glass injection  tee,
            deactivated (Supelco, Catalog no.   2-3665M).

                  4.1.3.3     Splitter   3   -   Restek   Y-shaped   fused-silica
            connector (Restek,  Catalog no.  20405).

            4.1.4 Column rinsing kit  (optional): Bonded-phase column  rinse kit
      (J&W Scientific, Catalog no. 430-3000 or equivalent).

            4.1.5 Microsyringes  -  100 /xL,  50 /xL,  10 JJ.L  (Hamilton  701  N or
      equivalent), and 50 /xL (Blunted, Hamilton 705SNR or  equivalent).

            4.1.6 Balances - Analytical, 0.0001 g.

            4.1.7 Volumetric flasks,  Class A - 10 ml to 1000 ml.


5.0   REAGENTS

      5.1   Reagent grade  inorganic chemicals shall be used  in all tests.  Unless
otherwise  indicated,  it  is  intended  that  all  reagents  shall conform  to the
specifications of the Committee  on Analytical Reagents of the American Chemical
Society, where  such  specifications are available.  Other  grades  may  be used,
provided it is  first  ascertained  that the chemicals are  of  sufficiently  high
purity to permit their use without affecting  the accuracy of the determinations.

      NOTE: Store  the  standard  solutions   (stock,  composite,   calibration,
            internal,  and surrogate)  at 4°C  in Teflon-sealed containers in the
            dark.   All standard solutions must be replaced after six months or
            sooner if routine QC (Sec. 8) indicates  a problem.
                                   8121 - 4                         Revision 0
                                                                September 1994

-------
      5.2   Solvents

            5.2.1 Hexane, C6H14  -  Pesticide  quality  or  equivalent.

            5.2.2 Acetone, CH3COCH3  -  Pesticide  quality or equivalent.

            5.2.3 Isooctane, (CH3)3CCH2CH(CH3)2 -  Pesticide quality or equivalent.

      5.3   Stock standard  solutions  (1000 mg/L):  Can be  prepared  from pure
standard materials or can be purchased as certified solutions.

            5.3.1 Prepare stock standard  solutions by accurately weighing about
      0.0100 g of pure compound.  Dissolve the compound in  isooctane or hexane
      and dilute to volume in a 10 ml volumetric flask.  If compound purity is
      96 percent  or greater,  the weight  can  be used without  correction  to
      calculate the concentration of the stock standard solution. Commercially
      prepared stock standard solutions can be used at any  concentration if they
      are certified by the manufacturer or by an independent source.

            5.3.2 For those compounds  which are not adequately soluble in hexane
      or isooctane,  mixtures of acetone and hexane are recommended.

      5.4   Composite  stock  standard: Can  be  prepared from  individual  stock
solutions.  For  composite  stock standards  containing  less than 25 components,
take exactly 1 mL of each individual stock solution at 1000 mg/L, add solvent,
and mix the solutions in  a 25 mL volumetric flask.  For  example, for a composite
containing 20  individual standards, the resulting concentration of each component
in the mixture,  after  the volume  is adjusted to 25  mL, will be 40 mg/L.   This
composite solution can be further diluted to obtain the desired concentrations.

      5.5   Calibration  standards should  be  prepared  at  a  minimum of  five
concentrations by dilution  of the composite stock  standard with isooctane or
hexane.    The concentrations   should  correspond  to   the  expected  range  of
concentrations found in real  samples and  should bracket the  linear range of the
detector.  A suggested list  of calibration  solution  standards is found in Table
7.

      5.6   Recommended  internal  standard: Make  a  solution  of 1000 mg/L  of
1,3,5-tnbromobenzene.   (Two  other  internal  standards, 2,5-dibromotoluene and
alpha,alpha'-dibromo-m-xylene,  are  suggested  if  matrix  interferences  are  a
problem.)  For spiking, dilute this solution to  50 ng//nL.   Use  a spiking volume
of 10 jiL/mL of extract.   The spiking concentration of the internal  standards
should be  kept constant  for all samples and calibration  standards.  Store the
internal standard spiking solutions at 4°C  in Teflon-sealed containers  in the
dark.

      5.7   Recommended surrogate standards:   Monitor  the  performance  of the
method using surrogate compounds.  Surrogate standards are  added to all samples,
method blanks, matrix  spikes,  and calibration  standards.    Make  a  solution  of
1000 mg/L of 1,4-dichloronaphthalene and  dilute  it to 100 ng/^l.  Use a spiking
volume of  100 yuL for  a  1  L  aqueous  sample.    If matrix   interferences  are  a
                                   8121 - 5                         Revision 0
                                                                September 1994

-------
problem,  two  alternative  surrogates  are:  alpha,  2,6-trichlorotoluene  or
2,3,4,5,6-pentachlorotoluene.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See the introductory material to this Chapter, Organic Analytes, Sec.
4.1.

      6.2   Extracts must  be stored  at  4 °C and  analyzed within 40  days  of
extraction.


7.0   PROCEDURE

      7.1   Extraction and Cleanup:

            7.1.1 Refer to Chapter Two and Method 3500 for guidance on choosing
      the  appropriate  extraction  procedure.   In  general,  water  samples  are
      extracted at a neutral, or as is, pH with methylene chloride, using either
      Method 3510  or  3520.   Solid samples  are extracted using  either Methods
      3540,  3541,  or  3550 with methylene  chloride/acetone  (1:1)  as  the
      extraction solvent.

            7.1.2 If required,  the samples may be cleaned up using Method 3620
      (Florisil)  and/or Method  3640  (Gel  Permeation  Chromatography).    See
      Chapter Two,  Sec.  2.3.2 and Method  3600 for  general  guidance on cleanup
      and method selection.  Method 3660  is used for sulfur removal.

            7.1.3 Prior to gas chromatographic analysis,  the extraction solvent
      must exchanged  into  hexane using the  Kuderna-Danish  concentration step
      found in any of the extraction  methods.  Any methylene chloride remaining
      in the extract will cause a very broad solvent peak.

      7.2   Gas Chromatographic Conditions:

            7.2.1 Retention  time  information for  each  of  the  analytes  is
      presented in Tables 3 and  4. The  recommended GC operating conditions are
      provided  in  Tables  5  and  6.   Figures  1,  2  and 3  illustrate  typical
      chromatography of the method analytes for both the  single column approach
      and the dual column approach when  operated at the conditions specified in
      Tables 5 and 6.

      7.3   Calibration:

            7.3.1 Prepare  calibration  standards  using the  procedures  in Sec.
      5.0.   Refer  to  Method   8000   for  proper calibration  procedures.  The
      procedure for internal or external calibration may be used.

            7.3.2 Refer to Method 8000 for the establishment of retention time
      windows.
                                   8121 - 6                         Revision 0
                                                                September 1994

-------
7.4   Gas chromatographic analysis:

      7.4.1 Method 8000 provides  instructions  on  the analysis sequence,
appropriate dilutions,  establishing daily  retention time  windows,  and
identification criteria.

      7.4.2 Automatic injections of 1 /xL are recommended.  Hand injections
of no more than 2  /uL  may be used if the analyst demonstrates quantitation
precision of < 10  percent  relative standard  deviation.  The solvent flush
technique may be used if the amount of solvent is kept at a minimum.   If
the  internal  standard  calibration technique is used,  add  10 pi  of  the
internal standard  to each mL of sample extract prior to injection.

      7.4.3 Tentative identification of an analyte occurs  when  a peak from
a sample extract falls within the  daily retention time window.

      7.4.4 Validation   of   gas   chromatographic   system   qualitative
performance: Use the midconcentration  standards  interspersed  throughout
the analysis sequence (Sec.  7.3)  to evaluate this  criterion.   If  any of
the standards  fall  outside their daily retention time windows,  the system
is out of control.  Determine  the cause of the problem and correct it (see
Sec. 7.5).

      7.4.5 Record the  volume  injected  to  the nearest  0.05 /iL and  the
resulting  peak  size  in  peak  height or  area  units.   Using  either  the
internal or the external  calibration procedure  (Method  8000), determine
the  identity  and   the  quantity of each  component peak  in  the  sample
chromatogram  which  corresponds to the compounds  used  for  calibration
purposes.  See Method 8000 for calculation equations.

      7.4.6 If the responses exceed the linear  range of the system,  dilute
the extract and reanalyze. Peak height measurements are recommended over
peak  area  integration  when  overlapping  peaks   cause   errors  in  area
integration.

      7.4.7 If partially overlapping or coeluting peaks are found,  change
columns  or try  a  GC/MS technique (see  Sec.  8.7 and  Method  8270).
Interferences  that prevent analyte identification and/or quantitation may
be removed by  the  cleanup techniques mentioned above.

      7.4.8 If the peak  response is less than 2.5 times the baseline noise
level, the validity of the quantitative result may be questionable.  The
analyst should consult  with the source of the sample to determine whether
further concentration of the sample is warranted.

7.5   Instrument Maintenance:

      7.5.1 Injection of sample extracts  from  waste sites often leaves a
high boiling residue  in: the injection port area, splitters when used,  and
the  injection  port  end  of  the  chromatographic  column.   This  residue
effects chromatography in many ways  (i.e.,  peak  tailing,  retention time
shifts, analyte degradation, etc.)  and, therefore, instrument maintenance


                             8121   - 7                         Revision 0
                                                          September 1994

-------
      is very important.  Residue buildup in a splitter may limit flow through
      one leg and therefore change the split ratios.  If this occurs during an
      analytical run, the quantitative data  may  be incorrect.   Proper cleanup
      techniques will minimize the problem and  instrument QC will indicate when
      instrument maintenance is required.

            7.5.2 Suggested chromatograph maintenance: Corrective measures may
      require any one or more  of the  following  remedial actions.  Also see Sec.
      7  in  Method  8000  for   additional  guidance  on corrective   action  for
      capillary columns and the injection port.

                  7.5.2.1     Splitter connections: For dual columns which are
            connected using a  press-fit  Y-shaped  glass splitter or a Y-shaped
            fused-silica connector, clean  and deactivate the splitter or replace
            with a cleaned and  deactivated splitter.   Break off the first few
            inches (up to one  foot)  of the  injection  port  side of  the column.
            Remove  the  columns  and  solvent  backflush   according  to  the
            manufacturer's instructions.   If these procedures fail to eliminate
            the degradation  problem,  it may be necessary to  deactivate the metal
            injector body and/or replace the columns.


8.0   QUALITY CONTROL

      8.1   Refer to Chapter One  and Method  8000  for  specific  quality control
procedures.   Quality control to validate  sample extraction  is covered in Method
3500 and in  the extraction method utilized.  If extract cleanup was performed,
follow the QC in Method 3600 and in the specific cleanup  method.

      8.2   Quality control required to evaluate the GC system operation is found
in Method 8000, Sec. 8.3.

      8.3   Calculate surrogate standard recoveries for all  samples,  blanks, and
spikes.   Determine  if  the  recovery  is  within limits (limits  established  by
performing QC procedures outlined in  Method 8000,  Sec. 8).   If the  recovery is
not within limits, the following are  required:

            8.3.1 Check  to  be  sure  there  are   no   errors  in  calculations,
      surrogate  solutions  and  internal   standards.    Also,  check  instrument
      performance.

            8.3.2 Recalculate  the data and/or  reanalyze the  extract if  any of
      the above checks reveal  a problem.

            8.3.3 Reextract and reanalyze the  sample  if none of the above are
      a problem, or flag the data as  "estimated concentrations".

      8.4   Data from systems that automatically identify target analytes on the
basis of  retention  time or retention  time  indices should be   reviewed  by  an
experienced  analyst before they are reported.

      8.5   When using the  internal standard  calibration technique,  an internal
standard peak  area check must be performed  on all samples.  The internal standard

                                   8121 - 8                         Revision 0
                                                                September 1994

-------
must be evaluated for acceptance by determining whether the measured area for the
internal standard deviates by more  than 50 percent from the average area for the
internal standard in the calibration standards.  When the internal  standard peak
area is outside that limit, all samples that fall outside the QC criteria must
be reanalyzed.

      8.6   Include a mid-concentration calibration standard after each  group of
20  samples   in  the   analysis  sequence.     The  response  factors  for  the
mid-concentration calibration  must be  within +  15 percent  of the average values
for the multiconcentration calibration.   When the response factors  fall outside
that  limit,   all  samples  analyzed after  that  mid-concentration  calibration
standard must be reanalyzed after performing instrument maintenance to correct
the usual source of the problem.   If  this  fails to correct the problem, a new
calibration curve must be established.

      8.7   GC/MS confirmation:

            8.7.1 GC/MS techniques  should  be judiciously  employed to support
      qualitative  identifications  made  with  this method.   Follow  the GC/MS
      operating requirements  specified  in  Method 8270.   Ensure  that there is
      sufficient concentration of  the  analyte(s)  to be confirmed,  in the extract
      for GC/MS analysis.

            8.7.2 When  available,   chemical  ionization  mass  spectra   may  be
      employed to aid  in the qualitative identification process.

            8.7.3 To confirm  an identification  of  a  compound,  the background
      corrected mass spectrum of the compound must be obtained from the sample
      extract  and  must  be compared  with  a mass  spectrum from  a  stock  or
      calibration standard analyzed under the same chromatographic conditions.
      At  least  25  ng  of  material  should  be  injected into  the  GC/MS.   The
      identification  criteria  specified  in  Method   8270  must   be  met  for
      qualitative confirmation.

                  8.7.3.1     Should   the   MS   procedure  fail    to  provide
            satisfactory   results,  additional   steps   may  be  taken  before
            reanalysis.  These steps may  include the use of alternate packed or
            capillary  GC columns or additional  sample cleanup.


9.0   METHOD  PERFORMANCE

      9.1   The MDL is defined in  Chapter One.   The  MDLs listed in  Table 1 were
obtained by using organic-free reagent water.   Details  on  how to determine MDLs
are given in  Chapter One.  The MDLs actually achieved in a given analysis will
vary since they depend on  instrument sensitivity and matrix effects.

      9.2   This  method  has  been  tested  in  a  single  laboratory  by  using
organic-free  reagent water, sandy  loam samples  and extracts which were spiked
with the  test compounds at one concentration.   Single-operator  precision and
method accuracy were found  to be related to the concentration of compound and the
type of matrix.


                                   8121  - 9                         Revision 0
                                                                September 1994

-------
      9.3   Single  laboratory  accuracy  data  were  obtained  for  chlorinated
hydrocarbons in a clay soil.  The spiking concentrations ranged from 500 to 5000
M9/kg,  depending on the  sensitivity  of the  analyte to the  electron capture
detector.  The spiking solution was mixed into the soil  during addition and then
immediatly transferred to the extraction device and immersed in the extraction
solvent.   The  spiked  sample  was then  extracted by  Method 3541  (Automated
Soxhlet).  The data represents a single determination.  Analysis was by capillary
column gas chromatography/electron capture detector following Method 8121 for the
chlorinated hydrocarbons.  These data  are listed  in Table 9 and were taken from
Reference 4.
10.0  REFERENCES

1.    Lopez-Avila,  V.,  N.S.  Dodhiwala,  and  J.  Milanes,   "Single  Laboratory
      Evaluation of Method 8120, Chlorinated Hydrocarbons", 1988, EPA Contract
      Numbers 68-03-3226 and 68-03-3511.

2.    Glazer, J.A., G.D. Foerst, G.D.  McKee, S.A.  Quave, and W.L. Budde, "Trace
      Analyses for Wastewaters," Environ.  Sci. and Techno!.  15:1426-1431, 1981.

3.    Lopez-Avila, V.;  Baldin,  E.;  Benedicto,  J; Milanes,  J.;  Beckert,  W.  F.
      "Application of Open-Tubular Columns to SW 846 GC Methods"; final report
      to  the U.S.  Environmental  Protection  Agency  on Contract  68-03-3511;
      Mid-Pacific Environmental Laboratory, Mountain View,  CA, 1990.

4.    Lopez-Avila, V. (Beckert, W., Project Officer), "Development of a Soxtec
      Extraction  Procedure for  Extracting Organic  Compounds from  Soils  and
      Sediments",  EPA  600/X-91/140,  US EPA, Environmental  Monitoring Systems
      Laboratory-Las Vegas, October 1991.
                                   8121  -  10                        Revision 0
                                                                September 1994

-------
                             TABLE  1
      METHOD DETECTION LIMITS FOR CHLORINATED HYDROCARBONS
                SINGLE COLUMN METHOD OF ANALYSIS
Compound name
Benzal chloride
Benzotrichloride
Benzyl chloride
2-Chloronaphthalene
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1 ,4-Dichlorobenzene
Hexachlorobenzene
Hexachlorobutadiene
a-Hexachlorocyclohexane (a-BHC)
/3-Hexachlorocyclohexane (/3-BHC)
y-Hexachlorocyclohexane (y-BHC)
5-Hexachlorocyclohexane (5-BHC)
Hexachl orocycl opentadi ene
Hexachloroethane
Pentachl orobenzene
1,2, 3, 4-Tetrachl orobenzene
1,2,4 , 5-Tetrachl orobenzene
1,2,3 , 5-Tetrachl orobenzene
1, 2, 4-Trichl orobenzene
1, 2, 3-Trichl orobenzene
1,3, 5 -Trichl orobenzene
CAS Reg. No.
98-87-3
98-07-7
100-44-7
91-58-7
95-50-1
541-73-1
106-46-1
118-74-1
87-68-3
319-84-6
319-85-7
58-89-9
319-86-8
77-47-4
67-72-1
608-93-5
634-66-2
95-94-2
634-90-2
120-82-1
87-61-6
108-70-3
MDLa
(ng/L)
2-5b
6.0
180
1,300
270
250
890
5.6
1.4
11
31
23
20
240
1.6
38
11
9.5
8.1
130
39
12
MDL is the method detection  limit for organic-free reagent water.  MDL
was determined from the  analysis of eight replicate aliquots processed
through the entire  analytical method  (extraction,  Florisil  cartridge
cleanup, and GC/ECD analysis).

      MDL - T/DC(n.1>B = .99,(s)

where t(n.1099) is  the  student's  t  value appropriate for  a  99  percent
confidence  interval and  a  standard  deviation with  n-1 degrees  of
freedom,  and  SD  is the standard  deviation of  the eight  replicate
measurements.

Estimated from the  instrument detection limit.
                              8121  -  11                         Revision 0
                                                            September 1994

-------
                                 TABLE 2

     ESTIMATED QUANTITATION  LIMIT  (EQL)  FACTORS  FOR VARIOUS MATRICES"

      Matrix                                                Factor

Ground water                                                     10

Low-concentration soil by ultrasonic extraction                 670
  with GPC cleanup

High-concentration soil and sludges by ultrasonic            10,000
  extraction

Waste not miscible with water                               100,000

8   EQL = [Method detection limit (see Table 1)] x [Factor found  in  this
    table].   For nonaqueous  samples,  the factor is  on  a wet-weight basis.
    Sample EQLs  are highly matrix-dependent.  The EQLs  listed  herein are
    provided for guidance and may not always be achievable.
                                   8121  - 12                        Revision 0
                                                                September 1994

-------
                                 TABLE 3
GAS CHROMATOGRAPHIC RETENTION TIMES FOR CHLORINATED  HYDROCARBONS:  SINGLE
                        COLUMN METHOD  OF ANALYSIS
Compound name
Retention time (min)
DB-2108DB-WAX"
Benzal chloride
Benzotrichloride
Benzyl chloride
2-Chloronaphthalene
1 , 2-Di chl orobenzene
1,3-Dichlorobenzene
1, 4- Di chl orobenzene
Hexachl orobenzene
Hexachl orobutadi ene
a-BHC
K-BHC
tf-BHC
Hexachl orocycl opentadi ene
Hexachl oroethane
Pent achl orobenzene
1,2,3,4-Tetrachlorobenzene
1,2,4, 5-Tetrachl orobenzene
1,2,3 , 5-Tetrachl orobenzene
1 ,2,4-Trichlorobenzene
1,2, 3 -Tri chl orobenzene
1, 3, 5-Tri chl orobenzene
6.86
7.85
4.59
13.45
4.44
3.66
3.80
19.23
5.77
25.54
24.07
26.16
8.86
3.35
14.86
11.90
10.18
10.18
6.86
8.14
5.45
15.91
15.44
10.37
23.75
9.58
7.73
8.49
29.16
9.98
33.84
54.30
33.79
c
8.13
23.75
21.17
17.81
17.50
13.74
16.00
10.37
Internal Standards

2,5-Dibromotoluene
1,3,5-Tribromobenzene
a,a'-Di bromo-meta-xyl ene

Surrogates
 9.55
11.68
18.43
(7,2,6-Trichlorotoluene        12.96
1,4-Dichloronaphthalene       17.43
2,3,4,5,6-Pentachlorotoluene  18.96
18.55
22.60
35.94
                  22.53
                  26.83
                  27.91
   GC  operating conditions:    30  m  x 0.53  mm  ID  DB-210  fused-silica
   capillary column; 1 jum film thickness;  carrier gas helium at 10 mL/min;
   makeup gas is nitrogen  at  40 mL/min; temperature  program  from 65°C to
   175°C (hold 20 minutes) at 4°C/min;  injector temperature 220°C; detector
   temperature 250°C.

   GC  operating conditions:    30  m  x 0.53  mm  ID  DB-WAX  fused-silica
   capillary column; 1 jiim film thickness;  carrier gas helium at 10 mL/min;
   makeup gas is nitrogen  at  40 mL/min; temperature  program  from 60°C to
   170°C (hold 30 minutes) at 4°C/min;  injector temperature 200°C; detector
   temperature 230°C.

   Compound decomposes on-column.
                                   8121  -  13
                                      Revision 0
                                  September 1994

-------
                                TABLE 4
            RETENTION TIMES OF THE CHLORINATED HYDROCARBONS'
                     DUAL COLUMN METHOD OF ANALYSIS
Compound
1,3-Dichlorobenzene
1,4-Dichlorobenzene
Benzyl chloride
1,2-Dichlorobenzene
Hexachloroethane
1,3,5-Trichlorobenzene
Benzal chloride
1,2,4-Trichlorobenzene
1,2,3-Trichlorobenzene
Hexachl orobutadi ene
Benzotrichloride
1,2,3 , 5-Tetrachl orobenzene
1,2,4,5-Tetrachlorobenzene
Hexachl orocycl opentad i ene
1,2, 3, 4-Tetrachl orobenzene
2-Chloronaphthalene
Pentachl orobenzene
a-BHC
Hexachl orobenzene
/3-BHC
7-BHC
5-BHC
DB-5
RT(min)
5.82
6.00
6.00
6.64
7.91
10.07
10.27
11.97
13.58
13.88
14.09
19.35
19.35
19.85
21.97
21.77
29.02
34.64
34.98
35.99
36.25
37.39
DB-1701
RT(min)
7.22
7.53
8.47
8.58
8.58
11.55
14.41
14.54
16.93
14.41
17.12
21.85
22.07
21.17
25.71
26.60
31.05
38.79
36.52
43.77
40.59
44.62
Internal Standard
1,3,5-Tribromobenzene                         11.83          13.34

Surrogate
1,4-Dichloronaphthalene                       15.42          17.71

"The  GC operating conditions  were  as follows:  30 m x  0.53  mm  ID  DB-5
(0.83-/xm   film thickness) and  30  m x 0.53  mm  ID  DB-1701  (1.0 /xm  film
thickness)  connected to an 8-in injection tee  (Supelco Inc.).  Temperature
program: 80°C (1.5 min hold) to 125°C (1 min hold) at 2°C/min then to 240°C
(2 min hold) at 5°C/m7n;  injector temperature 250°C;  detector  temperature
320°C; helium  carrier gas 6 mL/min; nitrogen makeup gas 20 mL/min.
                                   8121  -  14                         Revision 0
                                                                September 1994

-------
                              TABLE 5
             GC OPERATING CONDITIONS FOR CHLOROHYDROCARBONS
                    SINGLE COLUMN METHOD OF ANALYSIS
Column  1:  DB-210    30  m  x 0.53  mm  ID  fused-silica  capillary  column
chemically bonded with trifluoropropyl methyl silicone

    Carrier gas  (He)     10 mL/min
    Column temperature:
                  Initial  temperature        65°C
                  Temperature program        65°C  to  175°C  at  4°C/min
                  Final  temperature          175°C, hold 20 minutes.
    Injector  temperature       220°C
    Detector  temperature       250°C
    Injection volume           1-2 pL


    Column 2:  DB-WAX   30  m x 0.53 mm  ID fused-silica capillary  column
    chemically bonded with  polyethylene  glycol

    Carrier gas  (He)     10 mL/min
    Column temperature:
                  Initial  temperature        60°C
                  Temperature program        60°C  to  170°C  at  4°C/min
                  Final  temperature          170°C, hold 30 minutes.
    Injector  temperature       200°C
    Detector  temperature       230°C
    Injection volume           1-2 tit
                                   8121  -  15                         Revision 0
                                                                September 1994

-------
Column 1:
Column 2:
                              TABLE 6
           GC OPERATING CONDITIONS FOR CHLORINATED HYDROCARBONS
                      DUAL  COLUMN  METHOD  OF  ANALYSIS
                  Type:  DB-1701 (J&W Scientific) or equivalent
                  Dimensions:  30 m x 0.53 mm ID
                  Film Thickness: 1.0 (/urn)
                  Type:  DB-5 (J&W Scientific) or equivalent
                  Dimensions:  30 m x 0.53 mm ID
                  Film Thickness: 0.83
Carrier gas flowrate (mL/min):  6 (Helium)

Makeup gas flowrate (mL/min):  20 (Nitrogen)

Temperature program:  80°C (1.5 min hold) to 125"C (1 min hold)  at 2°C/min
then to 240°C (2 min hold) at 5eC/min.

Injector temperature:  250°C

Detector temperature:  320eC

Injection volume:  2 jiL

Solvent:  Hexane

Type of injector:  Flash vaporization

Detector type:  Dual ECD

Range:  10

Attenuation:   32 (DB-1701)/32 (DB-5)

Type of splitter:  Supelco 8-in  injection tee
                                  8121  -  16                         Revision 0
                                                                September 1994

-------
                                 TABLE 7

         SUGGESTED CONCENTRATIONS FOR THE CALIBRATION SOLUTIONS8


                              Concentration (ng//zL)
Benzal chloride
Benzotrichloride
Benzyl chloride
2-Chloronaphthalene
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1, 4- Dichl orobenzene
Hexachlorobenzene
Hexachl orobutadi ene
a-BHC
j8-BHC
7-BHC
6-BHC
Hexachl orocycl opentadi ene
Hexachl oroethane
Pentachl orobenzene
1,2,3,4-Tetrachlorobenzene
1 , 2 , 4, 5-Tetrachl orobenzene
1,2,3 , 5-Tetrachl orobenzene
1, 2, 4-Trichl orobenzene
1 ,2, 3 -Trichl orobenzene
1, 3, 5-Trichl orobenzene
0.1
0.1
0.1
2.0
1.0
1.0
1.0
0.01
0.01
0.1
0.1
0.1
0.1
0.01
0.01
0.01
0.1
0.1
0.1
0.1
0.1
0.1
0.2
0.2
0.2
4.0
2.0
2.0
2.0
0.02
0.02
0.2
0.2
0.2
0.2
0.02
0.02
0.02
0.2
0.2
0.2
0.2
0.2
0.2
0.5
0.5
0.5
10
5.0
5.0
5.0
0.05
0.05
0.5
0.5
0.5
0.5
0.05
0.05
0.05
0.5
0.5
0.5
0.5
0.5
0.5
0.8
0.8
0.8
16
8.0
8.0
8.0
0.08
0.08
0.8
0.8
0.8
0.8
0.08
0.08
0.08
0.8
0.8
0.8
0.8
0.8
0.8
1.0
1.0
1.0
20
10
10
10
0.1
0.1
1.0
1.0
1.0
1.0
0.1
0.1
0.1
1.0
1.0
1.0
1.0
1.0
1.0
Surrogates

a,2,6-Trichlorotoluene        0.02   0.05    0.1      0.15    0.2
1,4-Dichloronaphthalene       0.2    0.5     1.0      1.5     2.0
2,3,4,5,6-Pentachlorotoluene  0.02   0.05    0.1      0.15    0.2
    One  or  more  internal  standards  should  be  spiked prior  to  GC/ECD
    analysis into all  calibration solutions.   The spike concentration  of
    the  internal  standards  should  be kept constant  for all  calibration
    solutions.
                                  8121  -  17
    Revision 0
September 1994

-------
                               TABLE 8

             ELUTION PATTERNS OF CHLORINATED HYDROCARBONS
FROM THE FLORISIL COLUMN BY ELUTION WITH PETROLEUM ETHER (FRACTION 1)
         AND 1:1  PETROLEUM  ETHER/DIETHYL ETHER  (FRACTION 2)
Compound
Benzal chlorided
Benzotrichloride
Benzyl chloride
2-Chloronaphthalene
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
Hexachl orobenzene
Hexachl orobutadi ene
a-BHC
j3-BHC
7-BHC
5-BHC
Hexachl orocycl opentadi ene
Hexachl oroethane
Pentachl orobenzene
1,2, 3, 4-Tetrachl orobenzene
1 , 2 , 4 , 5-Tetrachl orobenzene8
1,2,3, 5-Tetrachl orobenzene6
1, 2, 4-Trichl orobenzene
1 , 2 , 3-Tri chl orobenzene
1 ,3 , 5-Tri chl orobenzene
Amount
(M9)
10
10
100
200
100
100
100
1.0
1.0
10
10
10
10
1.0
1.0
1.0
10
10
10
10
10
10
Recovery
Fraction 1"
0
0
82
115
102
103
104
116
101




93
100
129
104
102
102
59
96
102
(percent)"
Fraction 2C
0
0
16






95
108
105
71









  Values given represent average values of duplicate experiments.

  Fraction 1  was eluted with 200 mL  petroleum ether.

  Fraction 2  was eluted  with 200 mL petroleum ether/diethyl  ether (1:1).

  This    compound  coelutes   with   1,2,4-trichlorobenzene;   separate
  experiments were performed with benzal- chloride to verify  that  this
  compound is not recovered from the  Florisil  cleanup in either fraction.

  This   pair  cannot  be  resolved on  the  DB-210 fused-silica  capillary
  columns.
                                 8121  -  18
    Revision 0
September 1994

-------
                                    TABLE  9
             SINGLE  LABORATORY  ACCURACY  DATA FOR THE EXTRACTION OF
         CHLORINATED HYDROCARBONS  FROM  SPIKED CLAY SOIL BY METHOD 3541
                             (AUTOMATED SOXHLET)8
Compound Name
Spike Level
% Recovery
                                                      DB-5
                                    DB-1701
1, 3 -Dichl orobenzene
1,2-Dichlorobenzene
Benzal chloride
Benzotrichloride
Hexachl orocycl opentadi ene
Pentachl orobenzene
alpha-BHC
delta-BHC
Hexachl orobenzene
5000
5000
500
500
500
500
500
500
500
b
94
61
48
30
76
89
86
84
39
77
66
53
32
73
94
b
88
a     The  operating  conditions  for the  automated Soxhlet  were  as  follows:
      immersion time 45 min; extraction time  45  min;  the  sample size  was 10 g
      clay soil, extraction solvent,  1:1 acetone/hexane.  No equilibration time
      following spiking.

b     Not able to determine because of interference.

Data taken from Reference 4.
                                   8121  -  19
                                      Revision 0
                                  September 1994

-------
1

30
                                  IT
                            14
                         JU
                                                    10
                                                           73
                             K>
             IS

         T1MI (mtoi)
20
30
Figure 1.   GC/ECD chromatogram of Method 8121  composite standard analyzed on a
           30  m x  0.53 mm  ID  DB-210  fused-sH1ca  capillary column.    GC
           operating conditions  are given 1n Section 7.4.   See Table  3 for
           compound Identification.
                                 8121 - 20
                                         Revision 0
                                     September  1994

-------
                              17
                                               11
                                               13
             10     15
20     25     30     36     40     45     50    55
Figure 2.   GC/ECD chromatogram of Method 8121  composite standard analyzed  on  a
            30  m x  0.53  mm ID  DB-UAX  fused-silica  capillary column.   GC
            operating conditions  are given in  Section  7.4.   See Table 3  for
            compound Identification.
                                   8121  -  21
                                          Revision 0
                                      September 1994

-------
                                      DB-S

                                                            it
            [
ik
                                      DB-1701
                         10
                         i
                        7    .11     14 II     I*  IS     17   M    It   II  II    to
 If
uu
                                                              JU
Figure 3.   GC/ECD chromatogram  of  chlorinated hydrocarbons analyzed  on a DB
            5/DB 1701 fused-silica,  open-tubular column pair.  The GC operating
            conditions were as follows:  30 m  x  0.53  mm ID DB  5 (0.83 /xm film
            thickness) and 30 m  x 0.53 mm ID  DB  1701  (1.0 Mm film thickness)
            connected to  an  8 in injection  tee  (Supelco  Inc.).   Temperature
            program:  80°C  (1.5 min hold)  to 125°C  (1 min  hold) at 2°C/nrin, then
            to 240°C  (2 min hold) at 5°C/m1n.
                                   8121  -  22
                        Revision 0
                    September  1994

-------
                                                METHOD  8121

CHLORINATED HYDROCARBONS  BY  GAS  CHROMATOGRAPHY:  CAPILLARY  COLUMN  TECHNIQUE
                                                7.1.1 Chooee approon**
                                                 extraction procedure
                                              7.12 Add appropriate spiting
                                               c^ii^>ufi4li to uiiijii prior
                                                to extraction procedure
                                                       I
                                               7.2 Exchange extrecoon
                                               eofvent to hexane dicing
                                                  K-D procedure
                                             7.2.1 Following concentration of
                                              methytene crtoride allow K-O
                                               apparatus to drain vx) cool
                                           7^.2 IOCTMM MmpwMur* o! hot
                                             w«Mr tath; «dd hcauw: *tt*ch
                                        Snyd«r column: olco* «pcw«tu« on wiMr
                                             twti: ooncflntniM; remove from
                                                  «MM»r twth; cool
                                                       I
                                            72.3 Remove column; ririM Imak
                                              e/xl jointo with hexine; exljtut
                                                   •xtmct volume
          7.3 Choote •ppropriate deejxjp
             technique, if neceteery;
         (looroul cteeriup w recommended
            Refer to Method 3620 or to
                Section 7.3.2
                                                                       7^.3 Tnvwfer extract to
                                                                      Teflon teemed icrew-cip
                                                                         v«l»; refrigerate
                                                7.2.3 Will further
                                                 proceuJnebe
                                                performed withJn
                                                  twodcyi?
             73.3 GPC
              cleanup
              required?
                         7.3.4 Elemental
                         tuttur removeJ
                           required?
                                       7.3.4 Refer to
                                       Method 3880,
                                       Section 7.3
7.33 Refer to
Method 3640
No
                                                  8121  -  23
                                                                                         Revision 0
                                                                                   September  1994

-------
    METHOD  8121
    (continued)
   7.2.3 Stopper concentrator
        and refrigerate
 7.4.1 Set column 1 conditions
  7.4.2 Set column 2 conditions
  7.5.1 Refer to Method 8000 for
  calibration techniques; select
  lowest point on calibration curve
            I
  7.5.2 Choose and perform
 internal or external calibration
    (refer to Method 8000)
   7.6.1 Add internal standard
         if necessary
7.6.2 Establish daily retention time
  windows, analysis sequence,
dilutions, and identification criteria
         8121  -  24
      Revision  0
September  1994

-------
                        METHOD 8121
                        (concluded)
          0
  7.6.3 Record sample volume
  Injected and resulting peak
           sizes
  7.6.4 Determine identity and
quantity of each component peak
 that corresponds to compound
      used for calibration
           7.6.5
         Does peak
       exceed working
          range of
          system?
7.6.5 Dilute extract reanalyze
 7.6.6 Compare standard and
   sample retention times;
     identify compounds
                            8121  -  25
                           Revision  0
                     September  1994

-------
00

-------
                                 METHOD 8141

               ORGANOPHOSPHORUS COMPOUNDS  BY GAS CHROMATOGRAPHY:
                          CAPILLARY COLUMN TECHNIQUE
1.0   SCOPE AND APPLICATION

      1.1   Method 8141 is a gas chromatographic (GC) method used to determine
the concentration of various organophosphorus compounds.  The following compounds
can be determined by this method:
      Compound Name
CAS No.8
Azinphos-methyl
Bolstar (Sulprofos)
Chlorpyrifos
Coumaphos
Demeton, -0 and -S
Diazinon
Dichlorvos
Dimethoate
Disulfoton
EPN
Ethoprop
Fensulfothion
Fenthion
Malathion
Merphos
Mevinphos
Monocrotophos
Naled
Parathion, ethyl
Parathion, methyl
Phorate
Ronnel
Sulfotep
TEPP
Stirophos (Tetrachlorovinphos)
Tokuthion (Protothiofos)
Trichloronate
86-50-0
35400-43-2
2921-88-2
56-72-4
8065-48-3
333-41-5
62-73-7
60-51-5
298-04-4
2104-64-5
13194-48-4
115-90-2
55-38-9
121-75-5
150-50-5
7786-34-7
6923-22-4
300-76-5
56-38-2
298-00-0
298-02-2
299-84-3
3689-24-5
21646-99-1
22248-79-9
34643-46-4
327-98-0
      a  Chemical  Abstract Services Registry Number.

     1.2   Table 1 lists  method detection  limits  (MDL)  for  each compound in a
water and a soil matrix.   Table 2 lists the  estimated quantitation limits (EQLs)
for other matrices.
                                   8141 - 1
                      Revision 0
                       July 1992

-------
     1.3   Analytical difficulties encountered  with  specific organophosphorus
compounds may include (but are not limited to) the following:

           1.3.1  Tetraethyl  pyrophosphate  (TEPP)  is an  unstable  diphosphate
     which  is  readily  hydrolyzed  in  water  and  is thermally  labile  (TEPP
     decomposes  at 170°C).    Care must  be  taken to  minimize loss during  GC
     analysis and  during  sample preparation.    Identification  of  bad  standard
     lots is difficult since the electron impact mass  spectrum of TEPP is nearly
     identical to  its major breakdown product, triethyl  phosphate.

           1.3.2  The water solubility  of dichlorvos is  10 g/L  at  20°C,  and
     recovery is poor from aqueous solution.

           1.3.3  Naled  is converted  to Dichlorvos on column by debromination.
     This  reaction  may   also  occur  during  sample  workup.   The  extent  of
     debromination will  depend on the nature of  the matrix being analyzed.  The
     analyst must  consider the potential for debromination when Naled is to be
     determined.

           1.3.4  Trichlorofon (not determined by this method)  rearranges and is
     dehydrochlorinated in acidic, neutral, or  basic media  to form dichlorvos
     and hydrochloric acid.   If this  method  is to be  used for the determination
     of organophosphates in  the presence of  Trichlorofon, the analyst should be
     aware  of  the  possibility  of  rearrangement to Dichlorvos  to  prevent
     misidentification.

           1.3.5  Demeton    is    a    mixture    of    two    compounds;
     0,0-Diethyl 0-[2-(ethylthio)ethyl]   phosphorothioate   (Demeton-0)   and
     0,0-Diethyl S-[2-(ethylthio)ethyl]  phosphorothioate (Demeton-S).  Standards
     for  the individual  isomers  are  no   longer  available  through  the  EPA
     repository, and two peaks will be observed  in all mixed Demeton standards.
     It is recommended that the early eluting compound (Demeton-S) be used for
     quantitation.

           1.3.6  Tributyl phosphorotrithioite (Merphos)  is  a single component
     compound that is readily  oxidized in the  environment and during storage to
     the  phosphorotrithioate.   The  analyst  may  observe  two peaks   in  the
     chromatograms of merphos standards.

     1.4   Recoveries for some additional organophosphorus compounds have been
determined for water.  They include:

     Azinphos-ethyl              HMPA
     Carbofenthion               Leptophos
     Chlorfenvinphos             Phosmet
     Dioxathion                  Phosphamidion
     Ethion                      Terbuphos
     Famphur                     TOCP

     As Method 8141 has not  been fully validated  for  the determination of these
compounds, the analyst must demonstrate recoveries of greater than  70 percent
with precision of  no  more than 15 percent  RSD  before Method  8141  is  used for
these or any additional  analytes.


                                   8141 - 2                         Revision 0
                                                                     July 1992

-------
     1.5   When Method  8141  is used  to  analyze unfamiliar  samples,  compound
identifications should  be supported  by  at  least  two additional  qualitative
techniques if  mass  spectroscopy is  not  employed.    Section  8.4 provides  gas
chromatograph/mass spectrometer (GC/MS) criteria appropriate for the qualitative
confirmation of compound identifications.

     1.6   This method  is restricted to use by or  under the  supervision  of
analysts experienced in the use of gas chromatography and in the interpretation
of chromatograms.

     1.7   The use of Gel Permeation  Cleanup (Method 3640) for  sample cleanup has
been demonstrated to yield recoveries of less  than 85  percent for many method
analytes and is therefore not recommended for use with this method.


2.0  SUMMARY OF METHOD

     2.1   Method 8141  provides gas chromatographic conditions for the detection
of ppb concentrations of organophosphorus compounds.   Prior to the use of this
method, appropriate sample preparation techniques must be used.   Water samples
are extracted at  a neutral pH  with  methylene chloride  as  a solvent  by using a
separatory funnel  (Method 3510) or a continuous liquid-liquid extractor (Method
3520).  Soxhlet extraction (Method 3540)  or ultrasonic extraction (Method 3550)
using methylene chloride/acetone (1:1) are used  for solid samples. Both neat and
diluted organic liquids (Method 3580,  Waste Dilution) may be analyzed by direct
injection.  Spiked samples are used  to verify  the  applicability of the chosen
extraction technique to each  new sample type.  A gas chromatograph with a flame
photometric  or nitrogen-phosphorus  detector  is  used  for  this  multiresidue
procedure.

     2.2   If interferences are encountered in the analysis, Method 8141 may also
be performed on extracts that have undergone cleanup using Method 3620 or Method
3660.


3.0  INTERFERENCES

     3.1   Refer  to Methods 3500, 3600,   and 8000.

     3.2   The use of Florisil cleanup materials (Method 3620) for some of the
compounds in this  method has been demonstrated to yield recoveries less than 85%
and is therefore not recommended for all  compounds.   Refer  to  Table 2 of Method
3620 for  recoveries of  organophosphorus compounds  as a  function  of Florisil
fractions.   Use of phosphorus or  halogen specific  detectors,  however,  often
obviates  the  necessity for cleanup  for  relatively clean  sample matrices.   If
particular circumstances demand the  use of an alternative  cleanup procedure, the
analyst must determine the elution profile and  demonstrate  that the recovery of
each analyte is no less  than 85%.

     3.3   Use  of a flame photometric   detector  in  the phosphorus  mode will
minimize  interferences from materials that do not contain  phosphorus.  Elemental
sulfur, however, may interfere with the determination of certain organophosphorus
compounds by flame photometric gas chromatography.  Sulfur  cleanup using Method
3660 may  alleviate this  interference.

                                   8141  -  3                         Revision 0
                                                                     July 1992

-------
     3.4   A  halogen  specific  detector  (i.e.  electrolytic  conductivity  or
microcoulometric) is very selective for the halogen containing compounds and may
be used  for  the determination of Chlorpyrifos,  Ronnel,  Coumaphos,  Tokuthion,
Trichloronate, Dichlorvos, EPN,  Naled, and Stirophos only.

     3.5   Please note in Table 3 that a few analytes coelute on certain columns.
Therefore, select  a  second  column  for  confirmation  where coelution of  the
analytes of interest does no.t occur.

     3.6   Method  interferences  may  be  caused  by contaminants  in  solvents,
reagents, glassware, and other sample processing  hardware  that lead to discrete
artifacts or elevated baselines  in gas chromatograms.   All  these materials must
be routinely demonstrated to  be  free  from  interferences under the conditions of
the analysis by analyzing reagent blanks (refer to Section 8.1).


4.0  APPARATUS AND MATERIALS

     4.1   Gas chromatograph

           4.1.1  Gas  chromatograph,  analytical  system  complete  with  gas
     chromatograph and all required accessories including syringes, analytical
     columns,  gases,  detector   and   data  system,  integrator  or  stripchart
     recorder.  A  data system or integrator is recommended for measuring peak
     areas and/or peak heights.

           4.1.2  Columns

                  4.1.2.1     Column  1  -  15 m  x 0.53 mm megabore  capillary
           column, 1.0 urn film thickness, DB-210 (J&W Scientific  or equivalent.)

                  4.1.2.2     Column  2  -  15 m  x 0.53 mm megabore  capillary
           column, 1.5 pm film thickness, SPB-608 (Supelco, Inc.  or equivalent.)

                  4.1.2.3     Column  3  -  15 m  x 0.53 mm megabore  capillary
           column, 1.0 pm film thickness, DB-5 (J&W Scientific or equivalent.)

           4.1.3  Detector -  These detectors have proven effective in analysis
     for all analytes listed  in  Table 1 and Section 1.4 and were  used to develop
     the accuracy  and  precision  statements  in  Section  9.0.

                  4.1.3.1     Nitrogen Phosphorus Detector (NPD) operated in the
           phosphorus  specific mode is recommended.

                  4.1.3.2     Flame  Photometric  Detector  (FPD)  operated  in the
           phosphorus  specific mode is recommended.

                  4.1.3.3     Halogen   specific   detectors     (electrolytic
           conductivity  or microcoulometric) may be used  if only halogenated or
           sulfur  analytes are to be  determined.
                                   8141  - 4                         Revision 0
                                                                     July 1992

-------
     4.2   Kuderna-Danish (K-D) apparatus:

           4.2.1  Concentrator tube - 10 ml graduated (Kontes K-570050-1025 or
     equivalent).   A ground glass  stopper is used to  prevent  evaporation of
     extracts.

           4.2.2  Evaporation flask - 500 mL (Kontes K-570001-500  or equivalent).
     Attach to concentrator tube with springs, clamps or equivalent.

           4.2.3  Snyder column  -  Three  ball  macro (Kontes  K-503000-0121  or
     equivalent).

           4.2.4  Snyder column  -  Two ball  micro  (Kontes  K-569001-0219  or
     equivalent).

           4.2.5  Springs -  1/2 inch (Kontes  K-662750 or equivalent).

     4.3   Vials - 10 ml, glass with Teflon lined screw-caps or crimp tops.

     4.4   Water bath - Heated with concentric ring cover, capable of temperature
control (± 2°C).   The bath  should be used  in  a hood.

     4.5   Balance - Analytical, capable of accurately weighing to the nearest
0.0001 g.

     4.6   Boiling   chips   -   Solvent  extracted  with  methylene   chloride,
approximately 10/40 mesh (silicon carbide or equivalent).


5.0  REAGENTS

     5.1   Reagent grade inorganic chemicals shall be used in all tests.  Unless
otherwise indicated,  it  is intended that all inorganic reagents shall conform to
the specifications  of the  Committee on  Analytical  Reagents of  the  American
Chemical Society, where  such specifications are available.   Other grades may be
used,  provided it is first  ascertained that the reagent  is of sufficiently high
purity to permit its use without lessening the accuracy of the determination.

     5.2   Hexane, C6HU  - Pesticide quality or equivalent.

     5.3   Acetone, CH3COCH3 -  Pesticide quality  or equivalent.

     5.4   Isooctane, C8H18  - Pesticide  quality or equivalent.

     5.5   Surrogate standards - The analyst should monitor the performance of
the extraction,  cleanup  (when used), and analytical  system and  the effectiveness
of  the method  in  dealing  with  each  sample  matrix by spiking  each  sample,
standard, and blank with one or two  surrogates (e.g. organophosphorus compounds
not expected to  be present  in the sample)  recommended to encompass the range of
the temperature  program  used in  this method.  Deuterated  analogs of analytes
should  not  be  used as  surrogates  for gas   chromatographic  analysis due  to
coelution problems.
                                   8141 - 5                         Revision 0
                                                                     July 1992

-------
     5.6   Stock  standard  solutions  -  Can be  prepared  from pure  standard
materials or  can  be purchased as certified  solutions.   Commercially prepared
stock standards can be used if they  are  verified  against  EPA standards.  If EPA
standards are not  available  for  verification,  then standards  certified by the
manufacturer  and  verified  against  a  standard  made  from pure  material  is
acceptable.

           5.6.1  Prepare  stock   standard   solutions  by  accurately  weighing
     0.0100 g  of pure  material.    Dissolve  the  material  in  hexane  or  other
     suitable  solvent  and dilute to  known  volume  in a  volumetric  flask.   If
     compound  purity  is certified  at 96%  or greater,  the weight can be  used
     without correction  to  calculate  the concentration  of the stock standard.
     Commercially prepared stock standards can be used at any concentration if
     the are certified by the manufacturer or by an independent source.

           5.6.2  Transfer the stock standard solutions into bottles with Teflon
     lined  screw-caps  or crimp  tops.   Store  at 4°C and  protect  from light.
     Stock  standard  solutions  should   be  checked  frequently for  signs  of
     degradation or evaporation,  especially just  prior to preparing calibration
     standards from them.

           5.6.3  Stock standard  solutions  must be  replaced after six months or
     sooner if comparison with check  standards indicates a problem.   All  stock
     standards must be stored  in a  freezer at 4°C.

     5.7   Calibration  standards  -  A minimum of five concentrations for each
analyte of interest should be prepared through dilution of the stock standards
with isooctane. One of the concentrations should be at a  concentration near, but
above, the MDL.  The remaining concentrations should correspond to the expected
range of concentrations found in  real  samples or should define  the working range
of the GC.  Calibration standards must be replaced after one to two months, or
sooner if comparison with check standards indicates a problem.

     5.8   Internal standards should only be used  on well characterized samples.
To use this  approach, the analyst  must  select  one or more  internal standards that
are similar in analytical  behavior  to the  compounds  of  interest.   The analyst
must further demonstrate  that  the measurement  of the internal standard is not
affected by method  or  matrix  interferences.   Because of these limitations, no
internal standard can be suggested  that  is applicable to all  samples.

           5.8.1  Prepare   calibration   standards   at   a  minimum   of   five
     concentrations for  each  analyte  of  interest as described  in Section 5.7.

           5.8.2  To each calibration standard, add a known constant amount of
     one or more internal standards,  and dilute to volume with hexane or other
     suitable solvent.

           5.8.3  Analyze each calibration standard according  to Section  7.0.


6.0  SAMPLE COLLECTION,  PRESERVATION, AND  HANDLING

     6.1   See  the introductory  material  to this chapter,  Organic Analytes,
Section 4.1.

                                    8141  - 6                         Revision 0
                                                                     July 1992

-------
     6.2   Extracts are to be refrigerated at 4°C and analyzed within 40 days of
extraction.

7.0  PROCEDURE

     7.1   Extraction

           7.1.1  Refer to Chapter Two for guidance on choosing the appropriate
     extraction procedure.  In general, water samples are extracted at a neutral
     pH  with methylene  chloride,  using either  Method 3510  or 3520.   Solid
     samples  are  extracted  using either Method  3540  or 3550  with  methylene
     chloride/acetone  (1:1)  as the extraction solvent.

           7.1.2  Prior to gas chromatographic analysis, the extraction solvent
     may be  exchanged to  hexane.   The  exchange  is performed during  the  K-D
     procedures listed in all of  the extraction  methods.   The  analyst  must
     ensure quantitative transfer of the extract concentrate.  Single laboratory
     data  indicates  that samples should not  be transferred with  100 percent
     hexane  during  sample workup as  the more water  soluble  organophosphorus
     compounds  may  be  lost.    This  transfer  is  best  accomplished  with  a
     hexane/acetone solvent  mixture.  The exchange is performed as follows:

                  7.1.2.1     Following  K-D  concentration  of  the  methylene
           chloride extract  to 1 ml using  the macro Snyder column,  allow the
           apparatus to cool and  drain for at least 10 minutes.

                  7.1.2.2     Momentarily remove  the  Snyder column,  add 50 mL
           of hexane/acetone solvent mixture,  a new glass bead or boiling chip,
           and attach  the micro  Snyder column.   Concentrate the extract using
           1 ml of hexane to prewet  the Snyder column.  Place the K-D apparatus
           on the  water bath  so that the concentrator tube  is partially immersed
           in the  hot water.   Adjust the  vertical position of the apparatus and
           the water  temperature,  as required,  to complete  concentration in
           5-10 minutes.  At the proper  rate  of  distillation  the balls of the
           column will actively chatter,  but the chambers will not flood.  When
           the apparent volume of liquid  reaches 1  ml, remove the K-D apparatus
           and allow it to drain  and cool for at least 10 minutes.

                  7.1.2.3     Remove the Snyder column and rinse the flask and
           its lower joint with 1-2  ml of hexane into the concentrator tube.  A
           5 ml syringe is recommended for this operation.  Adjust the extract
           volume  to  10  ml.    Stopper  the  concentrator  tube  and  store
           refrigerated  at 4°C   if  further  processing  will  not be  performed
           immediately.  If the extract will be stored longer than two days, it
           should be  transferred to a vial with  a Teflon lined  screw-cap or
           crimp  top.   Proceed  with  gas chromatographic analysis  if further
           cleanup is  not required.

     7.2   Gas  chromatographic  conditions  (recommended):     Three  megabore
capillary columns  are included for analysis of organophosphates by this method.
Column 1  (DB-210  or  equivalent) and  Column  2  (SPB-608 or  equivalent)  are
recommended if a large  number of organophosphorus analytes are to be determined.
If the  superior resolution offered  by Column 1 and  Column 2  is not required,
Column 3 (DB-5 or equivalent) may be used.

                                   8141  - 7                         Revision 0
                                                                     July 1992

-------
           7.2.1  Columns 1 and 2

           Carrier gas (He) flow rate »   5 mL/min
           Initial temperature      =     50°C,  hold for 1  minute
           Temperature program      -     50°C to 140°C  at  50C/min, hold for 10
                                          minutes,  followed  by 140°C to 240°C
                                          at 10°C/min,  hold for 10 minutes (or
                                          a sufficient amount of time for last
                                          compound to elute).

           7.2.2'  Column  3

           Carrier gas (He) flow rate =   5 mL/min
           Initial temperature      -     130°C,  hold for 3 minutes
           Temperature program      =     130°C  to 180°C at 5°C/min, hold for 10
                                          minutes,  followed  by 180°C to 250°C
                                          at 2°C/min, hold  for  15 minutes (or a
                                          sufficient amount  of time  for  last
                                          compound to elute).

           7.2.3  Retention times for  all  analytes  on each column are presented
     in Table 3.  The analyst should note that several  method analytes coelute
     on column 3.

     7.3   Calibration:  Refer to Method 8000 for proper calibration techniques.
Use Table 1 and  especially Table 2 for guidance on selecting the lowest point on
the calibration curve.

           7.3.1  The procedure for internal  or external calibration may be used.
     Refer to Method 8000  for a description of each of these procedures.

           7.3.2  If cleanup  is  performed on the  samples, the analyst  should
     process  a  series of standards  through the  cleanup  procedure  and  then
     analyze  the  samples by GC.   This  will confirm elution patterns and the
     absence of interferences from the reagents.

     7.4   Gas chromatographic analysis

           7.4.1  Refer to Method 8000.   If the internal standard calibration
     technique  is used,  add  10 jtiL of internal standard to the  sample prior to
     injection.

           7.4.2  Method  8000 provides  instructions  on  the analysis sequence,
     appropriate  dilutions,   establishing daily  retention  time  windows,  and
     identification  criteria.

           7.4.3  For megabore capillary columns, automatic injections of 1 Mi-
     are recommended.  Hand injections of no more  than 2 pi may be used if the
     analyst  demonstrates quantitation   precision  of   <;  10 percent  relative
     standard deviation.   The solvent flush  technique may be  used  if the amount
     of solvent is kept  at a minimum.

           7.4.4  Examples  of  chromatograms   for  various   organophosphorus
     compounds  are shown  in Figures 1 through 4.

                                   8141 - 8                         Revision 0
                                                                     July 1992

-------
           7.4.5  Record the sample volume injected to the nearest  0.05  p.1 and
     the resulting peak sizes (in area units or peak heights).

           7.4.6  Using either the internal or  external  calibration  procedure
     (Method 8000), determine the identity and quantity of each component peak
     in the  sample chromatogram which  corresponds to the compounds  used for
     calibration purposes.  See Method 8000 for calculation equations.

           7.4.7  If peak  detection  and  identification  is  prevented  by  the
     presence of interferences, further cleanup is required.   Before using any
     cleanup  procedure,  the  analyst  must process  a  series of  calibration
     standards  through the  procedure  to  establish  elution  patterns  and  to
     determine recovery of target compounds.  The absence of interference from
     reagents must be  demonstrated  by routine  processing of reagent  blanks
     through the chosen cleanup procedure.

           7.4.8  Naled has been reported to be converted to DDVP on some columns
     by debromination.   If this  process  is demonstrated  on the GC  system that
     is used for analysis, clean  the injector  and break  off  several  inches of
     a megabore  column or change the  glass wool  of a packed  column  prior to
     analyzing samples.   If subsequent injections of Naled give DDVP,  report
     Naled  as  DDVP, but,  in  this instance,  both Naled  and  DDVP  may  not  be
     reported in the same sample.

     7.5   Cleanup

           7.5.1  Proceed with Florisil column Cleanup (Method 3620),  followed
     by, if  necessary, Sulfur Cleanup  (Method  3660),  using  the 10 mL  hexane
     extracts obtained  from Section 7.1.2.3.

           NOTE:  The use of Gel Permeation (Method 3640)  for sample cleanup has
                  been demonstrated to  yield recoveries of less than 85 percent
                  for many method analytes  and  is  therefore not recommended for
                  use with this method.

           7.5.2  Following cleanup,  the extracts should be analyzed by GC, as
     described in  the  previous Sections and in Method 8000.


8.0  QUALITY CONTROL

     8.1   Refer to Chapter One for specific quality control procedures.   Include
a mid-level  check  standard  after each group  of  10  samples   in the  analysis
sequence.   Quality  control to  validate sample  extraction is  covered in Method
3500 and in the extraction method utilized.  If extract cleanup was performed,
follow the QC in Method 3600 and in the specific cleanup method.

     8.2   Procedures to check the  GC system operation are found in Method 8000,
Section 8.6.

     8.3   GC/MS confirmation

           8.3.1  GC/MS techniques  should  be  judiciously  employed  to  support
     qualitative  identifications made with this method.   Follow the GC/MS

                                   8141 -  9                         Revision 0
                                                                     July 1992

-------
     operating requirements specified in Method 8270.

           8.3.2  When  available,  chemical  ionization  mass  spectra  may  be
     employed to aid in the qualitative identification process.

           8.3.3  To confirm an identification of a  compound, the background-
     corrected mass spectrum of the compound  must be obtained from the sample
     extract  and must  be  compared  with  a  mass  spectrum  from  a  stock  or
     calibration standard analyzed under  the  same chromatographic conditions.
     At  least 25  ng  of  material  should  be   injected  into the  GC/MS.    The
     following criteria must be met for qualitative confirmation:

           The  molecular  ion  and  all  other  ions  present  above 20  percent
     relative abundance in the mass spectrum of the standard must be present in
     the mass  spectrum of  the sample with agreement to ±  20  percent.    For
     example, if  the  relative abundance of an  ion  is 30 percent  in the  mass
     spectrum of the standard,  the allowable limits for the relative abundances
     of that ion in the mass spectrum  for  the  sample would be 20 to 40 percent.

           The retention  time of the  compound  in the sample must be within six
     seconds  of the  retention time  for  the  same  compound  in  the  standard
     solution.

           Compounds that  have very  similar  mass  spectra  can   be explicitly
     identified by GC/MS only on the  basis of retention time data.

           8.3.4  Where available, chemical  ionization  mass spectra  may  be
     employed to aid in the qualitative  identification process  because of the
     extensive  fragmentation  of  organophosphorus  compounds  during  electron
     impact MS processes.

           8.3.5  Should  the MS procedure  fail  to provide satisfactory results,
     additional steps may be taken before reanalysis.  These steps may include
     the use  of  alternate packed or capillary GC  columns or additional sample
     cleanup.


9.0  METHOD PERFORMANCE

     9.1   Estimated MDLs and associated chromatographic conditions for water and
clean soil  (uncontaminated with synthetic organics)  are listed in Table 1.  As
detection limit will vary with the particular matrix to be analyzed,  guidance for
estimating EQLs is given  in Table 2.

     9.2   Single operator  accuracy and precision studies  have  been conducted
with spiked water and soil  samples.  The results of these studies are presented
in Tables 4-7.
10.0 REFERENCES

1.   Taylor, V.;  Hickey,  D.M.;  Marsden,  P.O.  "Single Laboratory Validation of
     EPA  Method  8140";  U.S. Environmental  Protection  Agency.  Environmental
     Monitoring  Systems  Laboratory. Office  of Research and  Development,  Las

                                   8141  -  10                         Revision 0
                                                                     July 1992

-------
Vegas, NV, 1987; EPA-600/4-87-009.

Pressley, T.A;  Longbottom,  J.E.  "The Determination  of  Organophosphorus
Pesticides  in  Industrial  and  Municipal  Wastewater:  Method 614";  U.S.
Environmental  Protection Agency.  Environmental  Monitoring  and  Support
Laboratory.  Cincinnati, OH,  1982; EPA-600/4-82-004.

"Analysis of  Volatile Hazardous  Substances  by 6C/MS: Pesticide  Methods
Evaluation";  Letter  Reports  6,  12A, and  14  to the  U.S.Environmental
Protection Agency on Contract 68-03-2697, 1982.

"Method 622,  Organophosphorus  Pesticides";  U.S. Environmental  Protection
Agency. Environmental Monitoring Systems  Laboratory. Cincinnati, OH 45268.

Chau,  A.S.Y.; Afghan,  B.K.  Analysis  of Pesticides  in  Water. Vol.  II;
"Chlorine and Phosphorus-Containing Pesticides"; CRC: Boca Raton, FL, 1982,
pp 91-113, 238.

Hild,  J.;  Schulte,  E;  Thier,  H.P.  "Separation  of  Organophosphorus
Pesticides    and   Their   Metabolites   on   Glass-Capillary   Columns";
Chromatographia, 1978, 11-17.

Luke,  M.A.;   Froberg,   J.E.;   Doose,  G.M.;   Masumoto,   H.T.   "Improved
Multiresidue   Gas   Chromatographic   Determination  of  Organophosphorus,
Organonitrogen,  and  Organohalogen  Pesticides  in  Produce,  Using  Flame
Photometric and Electrolytic Conductivity Detectors";  vh. Assoc. Off. Anal.
Chem.  1981,  1187. 64.

Sherma,  J.;   Berzoa,  M.  "Analysis  of  Pesticide  Residues  in Human  and
Environmental  Samples";  U.S.  Environmental  Protection  Agency.  Research
Triangle  Park,  NC;  EPA-600/8-80-038.

Desmarchelier,  J.M.;  Wustner,  D.A.;  Fukuto,  T.R.  "Mass  Spectra  of
Organophosphorus Esters  and Their Alteration  Products";  Residue Reviews.
1974,  pp  63,  77.
                              8141 - 11                        Revision 0
                                                                July 1992

-------
                                   TABLE 1.
                 METHOD  DETECTION  LIMITS IN A WATER AND  A  SOIL
                   MATRIX  USING  A  FLAME  PHOTOMETRIC DETECTOR
Compound
   Reagent
Water (3510)8
   (M9/L)
                                                            Soil (3540)
Azinphos-methyl
Bolstar (Sulprofos)
Chlorpyrifos
Coumaphos
Demeton, -0, -S
Diazinon
Dichlorvos
Dimethoate
Disulfoton
EPN
Ethoprop
Fensulfothion
Fenthion
Malathion
Merphos
Mevinphos
Naled
Parathion-ethyl
Parathion-methyl
Phorate
Ronnel
Sulfotep
TEPPC
Tetrachlorovinphos
Tokuthion (Protothiofos)0
Trichloronatec
0.10
0.07
0.07
0.20
0.12
0.20
0.80
0.26
0.07
0.04
0.20
0.08
0.08
0.11
0.20
0.50
0.50
0.06
0.12
0.04
0.07
0.07
0.80
0.80
0.07
0.80
5.0
3.5
5.0
10.0
6.0
10.0
40.0
13.0
3.5
2.0
10.0
4.0
5.0
5.5
10.0
25.0
25.0
3.0
6.0
2.0
3.5
3.5
40.0
40.0
5.5
40.0
      a  Sample extracted using Method  3510,  Separatory Funnel  Liquid-Liquid
         Extraction.

      b  Sample extracted using Method 3540, Soxhlet Extraction.

      c  Purity of  these  standards  not established by  the  EPA  Pesticides and
         Industrial Chemicals Repository, RTP, NC.
                                   8141  -  12
                                Revision 0
                                 July 1992

-------
                                TABLE  2.
             DETERMINATION  OF  ESTIMATED QUANTITATION  LIMITS
                       (EQL)  FOR VARIOUS  MATRICES8
Matrix                                                            Factor6
Ground water (Methods 3510 or 3520)                                 10
Low-concentration soil by Soxhlet and no cleanup                    10°
Low-concentration soil by ultrasonic extraction with GPC cleanup     6.7C
High-concentration soil and sludges by ultrasonic extraction       500C
Non-water miscible waste (Method 3580)                            1000C
Sample EQLs are highly matrix dependent.  The  EQLs listed herein are provided
for guidance and may not always be achievable.

EQL =  [Method detection limit  (Table 1)] X  [Factor (Table 2)].  For non-
aqueous samples, the factor  is on a wet-weight basis.

Multiply this factor times the soil MDL.
                                8141 - 13                        Revision 0
                                                                  July  1992

-------
                TABLE 3.
RETENTION TIMES FOR METHOD 8141 ANALYTES

Compound
TEPP
Dichlorvos
Mevinphos
Demeton, -0 and -S
Ethoprop
Naled
Phorate
Monocrotophos
Sul f otep
Dimethoate
Disulfoton
Diazinon
Merphos
Ronnel
Chlorpyrifos
Malathion
Parathion, methyl
Parathion, ethyl
Trichloronate
Tetrachlorovinphos
Tokuthion (Protothiofos)
Fensulfothion
Bolstar^Sulprofos)
Famphur
EPN
Azinphos-methyl
Fenthion
Coumaphos
*Method 8141 has not been fully val
Initial temperature
Initial time
Program 1 rate
Program 1 final temperature
Program 1 hold
Program 2 rate
Program 2 final temperature
Program 2 hold
Capil
DBS
6.44
9.63
14.178
18.31
18.618
19.01
19.94
20.04
20.11
20.636
23.71
24.27
26.82
29.23
31.17
31.72
31.84
31.85
32.19
34.65
34.67
35.85
36.34
36.40
37.80
38.342
38.83
39.83
larv Column
SPB608
5.12
7.91
12.88
15.90
16.48
17.40
17.52
20.11
18.02
20.18
19.96
20.02
21.73
22.98
26.88
28.78
23.71
27.62
28.41
32.99
24.58
35.20
35.08
36.93
36.71
38.04
29.45
38.87

DB210
10.66
12.79
18.44
17.24
18.67
19.35
18.19
31.42
19.58
27.96
20.66
19.68
32.44
23.19
25.18
32.58
32.17
33.39
29.95
33.68
39.913
36.80
37.55
37.86
36.74
37.24
28.86
39.47
idated for Famphur.
130°C
3 minutes
5°C/niin
180°C
10 minutes
2°C/min
250°C
15 minutes
50°C
1 minute
5°C/min
140°C
10 minutes
10°C/min
240°C
10 minutes
50°C
1 minute
5°C/min
140°C
10 minutes
10°C/min
240°C
10 minutes
                8141 - 14
Revision 0
 July 1992

-------
                               TABLE 4.
    RECOVERY  OF 27  ORGANOPHOSPHATES  BY  SEPARATORY  FUNNEL  EXTRACTION
Compound
Azinphos-methyl
Bolstar
Chlorpyrifos
Coumaphos
Demeton
Diazinon
Dichlorvos
Dimethoate
Disulfoton
EPN
Ethoprop
Fensulfonthion
Fenthion
Malathion
Merphos
Mevinphos
Monocrotophos
Naled
Parathion, ethyl
Parathion, methyl
Phorate
Ronnel
Sulfotep
TEPP
Tetrachlorvinphos
Tokuthion
Trichloroate
Low
126
134
7
103
33
136
80
NR
48
113
82
84
NR
127
NR
NR
NR
NR
101
NR
94
67
87
96
79
NR
NR
Medium
143 + 8
141 + 8
89 + 6
90 + 6
67 + 11
121 + 9.5
79 + 11
47 + 3
92 + 7
125 + 9
90 + 6
82 + 12
48 + 10
92 + 6
79
NR
18 + 4
NR
94 + 5
46 + 4
77 + 6
97 + 5
85 + 4
55 + 72
90 + 7
45 + 3
35
High
101
101
86
96
74
82
72
101
84
97
80
96
89
86
81
55
NR
NR
86
44
73
87
83
63
80
90
94
NR = Not recovered.
                            8141  -  15
Revision 0
 July 1992

-------
                               TABLE  5.
   RECOVERY OF 27 ORGANOPHOSPHATES BY CONTINUOUS LIQUID EXTRACTION
Compound
Azinphos-methyl
Bolstar
Chlorpyrifos
Coumaphos
Demeton
Diazinon
Dichlorvos
Dimethoate
Disulfoton
EPN
Ethoprop
Famphur
Fensulfonthion
Fenthion
Malathion
Merphos
Mevinphos
Monocrotophos
Naled
Parathion, ethyl
Parathion, methyl
Phorate
Ronnel
Sulfotep
TEPP
Tetrachlorvinphos
Tokuthion
Trichloroate
Low
NR
NR
13
94
38
NR
81
NR
94
NR
39
--
90
8
105
NR
NR
NR
NR
106
NR
84
82
40
39
56
132
NR
Medium
129
126
82 + 4
79 + 1
23 + 3
128 + 37
32 + 1
10 + 8
69 + 5
104 + 18
76 + 2
63 + 15
67 + 26
32 + 2
87 + 4
80
87
30
NR
81 + 1
50 + 30
63+3
83 + 7
77 + 1
18 + 7
70 + 14
32 + 14
NR
High
122
128
88
89
41
118
74
102
81
119
83
--
90
86
86
79
49
1
74
87
43
74
89
85
70
83
90
21
NR = Not recovered.
                             8141  -  16
Revision 0
 July 1992

-------
                               TABLE 6.
        RECOVERY OF 27 ORGANOPHOSPHATES BY SOXHLET EXTRACTION
Compound
Azinphos -methyl
Bolstar
Chlorpyrifos
Coumaphos
Demeton
Diazinon
Dichlorvos
Dimethoate
Disulfoton
EPN
Ethoprop
Fensulfonthion
Fenthion
Malathion
Merphos
Mevinphos
Monocrotophos
Naled
Parathion., ethyl
Parathion, methyl
Phorate
Ronnel
Sulfotep
TEPP
Tetrachlorvinphos
Tokuthion
Trichloroate
Low
156
102
NR
93
169
87
84
NR
78
114
65
72
NR
100
62
NR
NR
NR
75
NR
75
NR
67
36
50
NR
56
Medium
110 + 6
103 + 15
66 + 17
89 + 11
64 + 6
96 + 3
39 + 21
48 + 7
78 + 6
93 + 8
70 + 7
81 + 18
43 + 7
81 + 8
53
71
NR
48
80 + 8
41+3
77 + 6
83 + 12
72 + 8
34 + 33
81 + 7
40 + 6
53
High
87
79
79
90
75
75
71
98
76
82
75
111
89
81
60
63
NR
NR
80
28
78
79
78
63
83
89
53
NR = Not recovered.
                             8141  -  17
Revision 0
 July 1992

-------
                               TABLE 7.
       RECOVERY OF  27 ORGANOPHOSPHATES  BY ULTRASONIC EXTRACTION
Compound
Azinphos-methyl
Bolstar
Chlorpyrifos
Coumaphos
Demeton
Diazinon
Dichlorvos
Dimethoate
Disulfoton
EPN
Ethoprop
Fensulfonthion
Fenthion
Malathion
Merphos
Mevinphos
Monocrotophos
Naled
Parathion, ethyl
Parathion, methyl
Phorate
Ronnel
Sulfotep
TEPP
Tetrachlorvinphos
Tokuthion
Trichloroate
Low
NR
NR
NR
NR
NR
NR
41
NR
30
14
19
NR
NR
55
NR
NR
NR
82
NR
63
NR
70
NR
43
NR
NR
NR
Medium
27 + 10
103 + 15
79 + 7
60
NR
90 + 14
13 + 9
67
44 + 22
86 + 38
34 + 26
37
35
67
71
NR
NR
40
74 + 13
NR
51+9
84+8
68 + 10
7
47 + 24
NR
NR
High
21
114
77
15
16
78
27
NR
69
105
35
2
84
31
155
23
NR
33
75
17
64
81
76
3
69
82
31
NR =. Not recovered.
                             8141  -  18                         Revision 0
                                                               July 1992

-------
                                      FIGURE 1.
      CHARACTERISTIC RESPONSE OF ORGANOPHOSPHATES ON OB210 WITH  FPD DETECTOR
300.00
250.00
200.00-
150.00
100.00
 50.00
  0.00




8
t
f- ...»-..
13579
1



L
i
m
S
°



(
i
(
<
i



1
1 1 1
Is
1 II

• . i

„

1
u.
1
I
1
$


1 1
2 8
1 1
e
s
v... jt










A
S

U

\
11 13 15 17 19 21 23 25 27 29 31 33 35 37 39 41 43 45
                                     8141  -  19
Revision 0
 July 1992

-------
                                        FIGURE  2.
       CHARACTERISTIC RESPONSE OF OR6ANOPHOSPHATES  ON DB210 WITH  NPD DETECTOR
300.00
250.00
200.00 -i
150.00-^
100.00-
 50.00-
  0.00 -J
III
               rf I I I | I I I | I lf|l fl|l»|MI|lll|ii||tli||ii|iM|iii|llTJ Tiipiinii|«TrjTi»|»t-i|1i»)

                5  7   9  11 13  15  17  19 21  23  25  27  29  31  33.  35 37  39 41  43  45
                                       8141 -  20
                                                    Revision  0
                                                     July  1992

-------
                                       FIGURE 3.
            CHROMATOGRAM OF ORGANOPHOSPHATES ON DB210 WITH FPD DETECTOR
300.00
                                 i
                                 Q
250.00
200.00
150.00
100.00
 50.00
                         II  .'
                                    \
                                                           
-------
                              FIGURE 4.
         CHROMATOGRAM OF ORGANOPHOSPHATES ON DB210 WITH NPO DETECTOR
300.00-,
250.00-
200.00-
150.00-
100.00-
50.00-^               .

    ;_JJL_jtL_
                                           |
                                              i
                                                    VI
                                                      to..
  0.00 •'-i •	i»" 11"!»"».' M...|...|.. .,,,.,,..,,,.,
       13  5  7  9  11  13 15  17 19 21  23 25 27 29 31 33 35 37 39  41 43  45
                             8141 - 22
                                                        Revision 0
                                                        July 1992

-------
                                          METHOD 8141
ORGANOPHOSPHORUS COMPOUNDS  BY  GAS  CHROMATOGRAPHY:  CAPILLARY  COLUMN  TECHNIQUE
                               7.1.1 Choose
                               appropriate
                            extraction procedure
                              (see Chapter 2)
                             7.1.2 Exchange
                            extraction solvent to
                              hexane during
                             K-D procedures
                                  1
                               7.2 Set gas
                             OvomatDgrapny
                               conditions
                               7.3 Refer to
                             Metfiod 8000 tor
                             proper calibration
                               techniques
                                                     7.3.2 Process a
                                                   series of stds through
                                                    cleanup procedure.
                                                     analyze by GC
                               7.4 Perform
                              QC analysis
                            (see Method 8000)
                                                    7.5.1 Cleanup using
                                                    erod 3620 and 3630
                                                      if necessary
                                          8141  -  23
Revision  0
 July  1992

-------
00

-------
                                 METHOD 8141A

               ORGANOPHOSPHORUS  COMPOUNDS  BY GAS  CHRQMATOGRAPHY:
                          CAPILLARY COLUMN TECHNIQUE
1.0   SCOPE AND APPLICATION

      1.1   Method 8141 is a capillary gas chromatographic (GC) method used to
determine the  concentration of organophosphorus  (OP)  compounds.   The  fused-
silica, open-tubular columns specified in this method offer improved resolution,
better  selectivity,  increased  sensitivity,  and  faster  analysis  than  packed
columns.  The compounds listed in  the  table below  can be determined by GC using
capillary columns  with  a  flame  photometric  detector  (FPD)  or  a  nitrogen-
phosphorus detector (NPD).  Triazine herbicides can also be determined with this
method when the NPD is used.  Although performance data  are presented for each
of the listed chemicals,  it  is unlikely that  all of them could be determined in
a  single  analysis.    This  limitation   results  because  the  chemical  and
chromatographic behavior of many  of these chemicals can  result in co-elution.
The analyst must select columns,  detectors and calibration  procedures for the
specific analytes of interest  in  a study.  Any listed chemical  is a potential
method interference when it is not a  target  analyte.
      Compound Name
                                  8141A - 1
CAS Registry No.
OP Pesticides
Aspon,b
Azinphos-methyl
Azinphos-ethyT
Bolstar (Sulprofos)
Carbophenothion8
Chlorfenvinphos8
Chlorpyrifos
Chlorpyrifos methyl8
Coumaphos
Crotoxyphos8
Demeton-0°
Demeton-Sc
Diazinon
Dichlorofenthion8
Dichlorvos (DDVP)
Dicrotophos8
Dimethoate
Dioxathion8'0
Disulfoton
EPN
Ethion0
Ethoprop
Famphur8
Fenitrothion8
Fensulfothion

3244-90-4
86-50-0
2642-71-9
35400-43-2
786-19-6
470-90-6
2921-88-2
5598-13-0
56-72-4
7700-17-6
8065-48-3
8065-48-3
333-41-5
97-17-6
62-73-7
141-66-2
60-51-5
78-34-2
298-04-4
2104-64-5
563-12-2
13194-48-4
52-85-7
122-14-5
115-90-2
              Revision 1
          September 1994

-------
      Compound Name
CAS Registry No.
Fonophos8
Fenthion
Leptophosa'd
Malathion
Merphos0
Mevinphos6
Monocrotophos
Naled
Parathion, ethyl
Parathion, methyl
Phorate
Phosmet8
Phosphamidon8
Ronnel
Stirophos (Tetrachlorovinphos)
Sulfotepp
TEPPd
Terbufos8
Thionazina-b (Zinophos)
Tokuthionb (Protothiofos)
Trichlorfon8
Trichloronate6
944-22-9
55-38-9
21609-90-5
121-75-5
150-50-5
7786-34-7
6923-22-4
300-76-5
56-38-2
298-00-0
298-02-2
732-11-6
13171-21-6
299-84-3
22248-79-9
3689-24-5
21646-99-1
13071-79-9
297-97-2
34643-46-4
52-68-6
327-98-0
Industrial Chemicals
      Hexamethylphosphoramide8 (HMPA)
      Tri-o-cresylphosphatea'd (TOCP)

Triazine Herbicides (NPD only)
      Atrazine8
      Simazine8
   680-31-9
    78-30-8
  1912-24-9
   122-34-9
      a     This analyte has been evaluated using a 30-m column only.
      b     Production discontinued in the U.S., standard not readily available,
      c     Standards may have multiple components because of oxidation.
      d     Compound is extremely toxic or neurotoxic.
      e     Adjacent major/minor peaks can be observed due to cis/trans isomers.
      1.2   A dual-column/dual-detector approach may be used for the analysis of
relatively clean extracts.   Two  15-  or 30-m x 0.53-mm  ID fused-silica,  open-
tubular columns of different polarities are  connected  to  an injection tee and
each is connected to  a detector.  Analysts  are cautioned regarding the use of a
dual column configuration when their instrument is  subject  to mechanical stress,
                                   8141A  -  2
              Revision  1
          September 1994

-------
when  many  samples  are  analyzed  over  a  short  time,  or  when  extracts  of
contaminated samples are analyzed.

      1.3   Two detectors can be used for the  listed OP chemicals. The FPD works
by measuring the emission of phosphorus- or sulfur-containing  species.  Detector
performance is  optimized by selecting the proper optical filter and adjusting the
hydrogen and air flows to the flame.   The  NPD is a flame ionization detector with
a rubidium  ceramic  flame tip which  enhances the response of phosphorus-  and
nitrogen-containing analytes.  The FPD is more sensitive and more selective, but
is a less common detector in environmental laboratories.

      1.4   Table 1  lists method detection limits  (MDLs) for the target analytes,
using 15-m  columns  and  FPD,  for water  and  soil  matrices.   Table 2  lists  the
estimated quantitation limits (EQLs) for other matrices.  MDLs  and EQLs using 30-
m columns will  be very similar to those obtained from 15-m columns.

      1.5   The use  of a  15-m column system has not been  fully validated for the
determination  of the  following  compounds.     The   analyst  must  demonstrate
chromatographic  resolution  of  all   analytes,  recoveries  of greater than  70
percent, with precision of no more than 15 percent RSD,  before data generated on
the 15-m column system can be reported for these, or any additional,  analytes:

      Azinphos-ethyl    Ethion      Phosmet
      Carbophenothion   Famphur     Phosphamidon
      Chlorfenvinphos   HMPA        Terbufos
      Dioxathion        Leptophos   TOCP

      1.6   When Method  8141  is used to analyze unfamiliar  samples,  compound
identifications should be supported by confirmatory analysis.   Sec. 8.0 provides
gas  chromatograph/mass   spectrometer  (GC/MS)   criteria   appropriate  for  the
qualitative confirmation of compound identifications.

      1.7   This method  is restricted to use  by, or  under the supervision of,
analysts  experienced  in  the use  of capillary  gas  chromatography  and  in  the
interpretation  of chromatograms.


2.0   SUMMARY OF METHOD

      2.1   Method 8141 provides gas  chromatographic conditions for the detection
of ppb concentrations of organophosphorus compounds.   Prior to the  use of this
method,  appropriate sample preparation techniques must be used.   Water samples
are extracted  at a  neutral  pH  with  methylene chloride   by using a  separatory
funnel  (Method  3510)  or  a  continuous  liquid-liquid  extractor  (Method  3520).
Soxhlet  extraction (Method 3540) or automated Soxhlet extraction (Method 3541)
using methylene chloride/acetone (1:1) are used for solid samples. Both neat and
diluted  organic liquids  (Method  3580, Waste Dilution)  may  be  analyzed by direct
injection.  Spiked samples are  used  to  verify the applicability of  the chosen
extraction technique to  each  new sample  type.  A  gas chromatograph with a flame
photometric  or nitrogen-phosphorus  detector  is used  for  this multiresidue
procedure.
                                  8141A  -  3                         Revision 1
                                                                September 1994

-------
      2.2   Organophosphorus esters and thioesters can hydrolyze under both acid
and  base    conditions.    Samples prepared  using acid  and  base  partitioning
procedures are not suitable for analysis by Method 8141.

      2.3   Ultrasonic Extraction (Method  3550)  is not  an  appropriate sample
preparation method  for Method  8141   and  should  not  be  used  because  of  the
potential  for destruction of target  analytes  during  the ultrasonic extraction
process.


3.0   INTERFERENCES

      3.1   Refer to Methods 3500, 3600, and 8000, as well as to Sec. 1.1.

      3.2   The use  of Florisil Cleanup  (Method 3620)  for some of the compounds
in this  method has been demonstrated to yield recoveries less than 85 percent and
is therefore not recommended  for all  compounds.  Refer to Table 2 of Method 3620
for recoveries of organophosphorus compounds.  Use of an FPD often eliminates the
need for  sample cleanup.    If  particular circumstances  demand  the  use  of an
alternative cleanup  procedure, the analyst must determine the elution profile and
demonstrate that the recovery of each analyte is not less than 85 percent.

      3.3   The  use of Gel  Permeation Cleanup (GPC)  (Method  3640)  for sample
cleanup has been demonstrated to  yield  recoveries of  less than  85 percent  for
many method analytes  because they elute before  bis-(2-ethylhexyl)  phthalate.
Method  3640  is   therefore  not  recommended  for use with this  method,  unless
analytes of  interest  are listed  in  Method 3640  or  are  demonstrated  to  give
greater than 85 percent recovery.

      3.4   Use  of  a  flame photometric detector  in  the  phosphorus  mode  will
minimize interferences from materials that do  not contain  phosphorus or sulfur.
Elemental    sulfur   will   interfere   with   the   determination  of  certain
organophosphorus compounds by flame  photometric gas chromatography.   If Method
3660 is  used for sulfur cleanup, only the  tetrabutylammonium (TBA)-sulfite option
should  be employed, since  copper and mercury may destroy  OP  pesticides.   The
stability of each analyte must be tested to ensure  that the recovery from  the
TBA-sulfite sulfur cleanup step is not less than 85 percent.

      3.5   A halogen-specific  detector (i.e.,  electrolytic conductivity  or
microcoulometry) is  very  selective for the halogen-containing compounds and may
be used for  the determination  of Chlorpyrifos,  Ronnel,  Coumaphos,  Tokuthion,
Trichloronate,   Dichlorvos,  EPN,   Naled,  and  Stirophos only.   Many  of  the  OP
pesticides may  also  be detected by the electron capture detector  (ECD); however,
the ECD is not  as specific as the NPD or FPD.  The ECD should only be used when
previous analyses have demonstrated that interferences  will  not adversely effect
quantitation,  and that the detector sensitivity is sufficient to meet regulatory
1imits.

      3.6   Certain analytes will coelute,  particularly on 15-m columns (Table
3).  If coelution  is  observed,  analysts should  (1) select  a  second  column of
different polarity for confirmation,  (2) use 30-m  x 0.53-mm columns, or (3)  use
0.25- or  0.32-mm ID  columns.   See  Figures 1 through  4 for combinations  of
compounds that do not coelute on  15-m columns.


                                  8141A - 4                        Revision 1
                                                                September 1994

-------
 3.7   The following pairs coeluted on the DB-5/DB-210 30-m column pair:

 DB-5  Terbufos/tri-o-cresyl  phosphate
       Naled/Simazine/Atrazine
       Dichlorofenthion/Demeton-0
       Trichloronate/Aspon
       Bolstar/Stirophos/Carbophenothion
       Phosphamidon/Crotoxyphos
       Fensulfothion/EPN

DB-210 Terbufos/tri-o-cresyl  phosphate
       Dichlorofenthion/Phosphamidon
       Chlorpyrifos, methyl/Parathion, methyl
       Chlorpyrifos/Parathion, ethyl
       Aspon/Fenthion
       Demeton-0/Dimethoate
       Leptophos/Azinphos-methyl
       EPN/Phosmet
       Famphur/Carbophenothi on

 See Table 4  for retention times  of these compounds on 30-m columns.

 3.8   Analytical  difficulties encountered for target analytes include:

       3.8.1  Tetraethyl  pyrophosphate  (TEPP)  is an  unstable  diphosphate
 which  is  readily  hydrolyzed in  water  and  is  thermally labile  (TEPP
 decomposes at  170°C).   Care  must  be taken  to  minimize  loss during  GC
 analysis and during sample preparation.   Identification  of bad  standard
 lots is difficult since the electron  impact  (El) mass spectrum of TEPP is
 nearly identical  to its major breakdown product,  triethyl phosphate.

       3.8.2  The water solubility of Dichlorvos (DDVP) is 10 g/L  at 20°C,
 and recovery is poor from aqueous solution.

       3.8.3  Naled  is  converted  to  Dichlorvos   (DDVP)  on  column   by
 debromination.   This reaction may  also  occur  during  sample workup.   The
 extent of debromination  will  depend on  the  nature of the  matrix being
 analyzed.  The  analyst  must consider  the potential for debromination when
 Naled is to  be  determined.

       3.8.4  Trichlorfon rearranges and  is dehydrochlorinated  in  acidic,
 neutral, or  basic media to form  Dichlorvos (DDVP)  and hydrochloric acid.
 If this method  is to  be  used for the determination of organophosphates in
 the  presence  of  Trichlorfon,   the' analyst  should  be  aware   of  the
 possibility  of  rearrangement  to  Dichlorvos to prevent misidentification.

       3.8.5  Demeton (Systox)   is  a  mixture of two  compounds;  0,0-diethyl
 0-[2-(ethylthio)ethyl]phosphorothioate (Demeton-0) and 0,0-diethyl S-[2-
 (ethylthio)ethyl]phosphorothioate (Demeton-S).  Two peaks are observed in
 all  the  chromatograms  corresponding  to  these   two  isomers.    It   is
 recommended   that  the  early  eluting  compound (Demeton-S)  be used  for
 quantitation.
                             8141A - 5                         Revision 1
                                                           September 1994

-------
            3.8.6 Dioxathion is a single-component  pesticide.  However, several
      extra peaks are observed in the chromatograms of standards.  These peaks
      appear  to be  the  result  of  spontaneous oxygen-sulfur  isomerization.
      Because  of  this,   Dioxathion  is  not  included  in  composite  standard
      mixtures.

            3.8.7 Merphos (tributyl phosphorotrithioite) is a single-component
      pesticide  that  is  readily oxidized to  its  phosphorotrithioate (Merphos
      oxone).   Chromatographic analysis of Merphos almost  always  results two
      peaks  (unoxidized  Merphos elutes  first).    As  the relative  amounts of
      oxidation  of  the  sample  and   the   standard   are  probably  different,
      quantitation based on the sum of both peaks may be most appropriate.

            3.8.8 Retention times of some analytes, particularly Monocrotophos,
      may increase with  increasing concentrations  in the  injector.   Analysts
      should check for retention time shifts in highly contaminated samples.

            3.8.9 Many  analytes  will   degrade  on   reactive   sites  in  the
      chromatographic system.   Analysts must ensure  that  injectors and splitters
      are  free  from  contamination  and are  silanized.    Columns   should  be
      installed and maintained properly.

            3.8.10      Performance of chromatographic systems will degrade with
      time.   Column resolution, analyte  breakdown and  baselines may be improved
      by column washing (Sec.  7).  Oxidation of columns  is not  reversible.

      3.9   Method interferences  may  be caused  by contaminants  in  solvents,
reagents, glassware,  and other sample processing hardware that lead to discrete
artifacts or elevated baselines in  gas chromatograms.  All these materials must
be routinely demonstrated to be free from interferences under the conditions of
the analysis by analyzing reagent blanks (Sec. 8.0).

      3.10  NP Detector interferences:    Triazine herbicides, such as Atrazine
and Simazine, and other nitrogen-containing compounds may interfere.


4.0   APPARATUS AND MATERIALS

      4.1   Gas  chromatograph:  An analytical  system  complete  with  a  gas
chromatograph  suitable  for on-column  or spl it/splitless  injection, and  all
required accessories, including  syringes, analytical  columns,  gases,  suitable
detector(s), and a  recording device.  The analyst should  select the detector for
the specific measurement application,  either the flame photometric detector or
the nitrogen-phosphorus detector.  A data'system for  measuring  peak  areas and
dual display of chromatograms  is highly recommended.

            4.1.1 Capillary Columns (0.53-mm,  0.32-mm, or 0.25-mm ID x 15-m or
      30-m length,  depending on the resolution required).   Columns of 0.53-mm ID
      are recommended for most  environmental  and waste  analysis applications.
      Dual-column,  single-injector analysis  requires columns of equal length and
      bore.   See Sec.  3.0 and  Figures 1  through 4 for guidance on selecting the
      proper length and diameter for the column(s).
                                   8141A  -  6                         Revision 1
                                                                September 1994

-------
            4.1.1.1     Column  1  -   15-  or  30-m  x  0.53-mm  wide-bore
      capillary column,  l.Q-/j,m film thickness,  chemically bonded with 50%
      trifluoropropyl polysiloxane,  50% methyl  polysiloxane  (DB-210), or
      equivalent.

            4.1.1.2     Column  2  -   15-  or  30-m  x  0.53-mm  wide-bore
      capillary column,  0.83-/zm  film thickness,   chemically bonded with
      35%  phenyl  methyl    polysiloxane   (DB-608,  SPB-608,  RTx-35),  or
      equivalent.

            4.1.1.3     Column  3  -   15-  or  30-m  x  0.53-mm  wide-bore
      capillary column,  1.0 p,m film thickness,   chemically  bonded with 5%
      phenyl  polysiloxane,  95% methyl  polysiloxane (DB-5,  SPB-5, RTx-5),
      or equivalent.

            4.1.1.4     Column 4 - 15- or 30-m x 0.53-mm ID fused-silica
      open-tubular column,  chemically bonded  with methyl   polysiloxane
      (DB-1, SPB-1, or equivalent), 1.0-/Lim or  1.5-/um film  thickness.

            4.1.1.5     (optional) Column  rinsing kit: Bonded-phase column
      rinse kit (J&W Scientific,  Catalog  no.  430-3000 or equivalent).

      4.1.2 Splitter: If a dual-column,  single-injector configuration is
used, the open tubular columns should be connected to one  of the following
splitters, or equivalent:

            4.1.2.1     Splitter  1  -  J&W Scientific  press-fit  Y-shaped
      glass 3-way union  splitter (J&W Scientific, Catalog no. 705-0733).

            4.1.2.2     Splitter 2  -  Supelco 8-in  glass  injection tee,
      deactivated (Supelco, Catalog no.  2-3665M).

            4.1.2.3     Splitter   3   -    Restek   Y-shaped   fused-silica
      connector (Restek, Catalog no.  20405).

      4.1.3 Injectors:

            4.1.3.1     Packed column, 1/4-in injector port  with hourglass
      liner are recommended for 0.53-mm column.   These  injector ports can
      be fitted with splitters (Sec.  4.0)  for dual-column analysis.

            4.1.3.2     Split/splitless capillary  injectors  operated in
      the split mode are required for 0.25-mm and 0.32-mm columns.

      4.1.4 Detectors:

            4.1.4.1     Flame  Photometric Detector (FPD)  operated in the
      phosphorus-specific  mode is recommended.

            4.1.4.2     Nitrogen-Phosphorus Detector (NPD) operated  in the
      phosphorus-specific  mode is less selective but can  detect  triazine
      herbicides.
                            8141A  - 7                         Revision 1
                                                          September 1994

-------
                  4.1.4.3     Halogen-Specific    Detectors     (electrolytic
            conductivity or  microcoulometry)  may be  used only  for  a limited
            number of halogenated or sulfur-containing analytes  (Sec. 3.0).

                  4.1.4.4     Electron-capture  detectors  may  be  used  for  a
            limited number of analytes (Sec. 3.0).

            4.1.5 Data system:

                  4.1.5.1     Data system capable of presenting chromatograms,
            retention time, and peak integration data is strongly recommended.

                  4.1.5.2     Use of a data system  that  allows storage  of raw
            chromatographic data is strongly recommended.


5.0   REAGENTS

      5.1   Solvents

            5.1.1 Isooctane,  (CH3)3CCH2CH(CH3)2 - Pesticide quality or equivalent.

            5.1.2 Hexane, C6H14  - Pesticide quality or equivalent.

            5.1.3 Acetone,  CH3COCH3  - Pesticide quality or equivalent.

            5.1.4 Tetrahydrofuran (THF),  C4H80 - Pesticide  quality or equivalent
      (for triazine standards only).

            5.1.5 Methyl tert-butyl-ether (MTBE), CH3Ot-C4H9 - Pesticide quality
      or equivalent (for triazine standards only).

      5.2   Stock standard solutions  (1000 mg/L):  Can  be prepared  from  pure
standard materials or can be  purchased as certified solutions.

            5.2.1 Prepare stock standard  solutions by accurately weighing about
      0.0100 g of pure compounds.  Dissolve the compounds  in  suitable mixtures
      of acetone and hexane and  dilute to  volume  in  a  10-mL  volumetric  flask.
      If compound  purity  is  96 percent  or greater,  the weight can be  used
      without correction to calculate the  concentration of the stock standard
      solution.  Commercially prepared stock standard solutions can be used at
      any concentration  if they are  certified by the  manufacturer or by  an
      independent source.

            5.2.2 Both Simazine  and Atrazine have low  solubilities  in hexane.
      If Simazine  and  Atrazine standards  are required, Atrazine  should  be
      dissolved in MTBE, and  Simazine should be dissolved in acetone/MTBE/THF
      (1:3:1).

            5.2.3 Composite stock standard:  This standard can be prepared  from
      individual  stock   solutions.   The analyst  must demonstrate  that  the
      individual analytes  and common  oxidation  products  are resolved  by  the
      chromatographic system.   For  composite   stock  standards  containing  less


                                  8141A  -  8                         Revision  1
                                                                September  1994

-------
      than 25 components,  take exactly 1 ml of each individual  stock solution at
      1000 mg/L, add solvent, and mix the solutions in a 25-mL  volumetric flask.
      For  example,  for a  composite containing  20  individual  standards,  the
      resulting concentration of each component in the mixture, after the volume
      is adjusted to  25 ml,  will  be 40 mg/L.  This  composite solution can be
      further diluted  to  obtain the desired  concentrations.   Composite stock
      standards containing more than 25 components are not recommended.

            5.2.4 Store the standard solutions (stock, composite, calibration,
      internal, and surrogate)  at 4°C in Teflon-sealed containers in the dark.
      All standard solutions should be replaced after two months, or sooner if
      routine  QC  (Sec.   8.0)   indicates  a  problem.    Standards  for  easily
      hydrolyzed chemicals including TEPP, Methyl  Parathion, and Merphos should
      be checked every 30 days.

            5.2.5 It  is recommended that  lots of  standards be subdivided  and
      stored  in small  vials.    Individual  vials  should be  used as  working
      standards to minimize the potential for contamination or  hydrolysis of the
      entire lot.

      5.3   Calibration standards   should be  prepared  at a  minimum  of  five
concentrations  by dilution of  the  composite  stock standard with  isooctane or
hexane.     The   concentrations  should  correspond  to  the expected  range  of
concentrations found in real  samples and should bracket the linear range of the
detector.  Organophosphorus calibration standards should be  replaced  after one
or two months,  or sooner  if  comparison with  check samples or historical  data
indicates that  there  is a problem.   Laboratories  may wish to  prepare separate
calibration solutions for the easily hydrolyzed standards identified  above.

      5.4   Internal standard:  Internal  standards should only  be used on well-
characterized samples by analysts experienced  in the technique. Use of internal
standards  is  complicated  by co-elution  of  some OP  pesticides  and  by  the
differences in detector response to dissimilar chemicals.

            5.4.1 FPD response for organophosphorus compounds is enhanced by the
      presence of sulfur atoms  bonded to the phosphorus atom.   It has not been
      established that a thiophosphate can be used as an internal  standard for
      an OP with a  different numbers  of  sulfur  atoms (e.g.,  phosphorothioates
      [P=S] as  an internal standard for phosphates  [P04]) or phosphorodithioates
      [P=S2]).

            5.4.2 If internal standards  are  to be  used, the analyst must select
      one or more internal  standards that are  similar in analytical behavior to
      the compounds  of interest.  The analyst  must further demonstrate that the
      measurement of the internal  standard is not affected by  method  or matrix
      interferences.

            5.4.3 When 15-m  columns are used,  it may  be difficult to  fully
      resolve internal standards from target analytes, method  interferences and
      matrix interferences.   The analyst must demonstrate that the measurement
      of  the   internal   standard   is   not    affected  by  method  or   matrix
      interferences.
                                  8141A  -  9                         Revision 1
                                                                September 1994

-------
            5.4.4 The following NPD internal standard has been used for a 30-m
      column pair.   Make  a solution of 1000 mg/L of  l-bromo-2-nitrobenzene.  For
      spiking, dilute this  solution to 5 mg/L.  Use  a  spiking volume of 10 /iL/mL
      of extract. The spiking concentration of  the  internal standards should be
      kept constant for  all  samples and  calibration  standards.   Since its FPD
      response is small,  l-bromo-2-nitrobenzene is not an appropriate internal
      standard for that detector.  No FPD  internal  standard is suggested.

      5.5   Surrogate standard spiking solutions - The analyst should monitor the
performance of the extraction, cleanup (when used), and analytical system, and
the effectiveness of the method in dealing with each sample matrix,  by spiking
each  sample,  standard,   and  blank   with  one   or   two  surrogates  (e.g.,
organophosphorus compounds  not expected  to be  present   in  the  sample).   If
multiple analytes are to be measured, two  surrogates (an early and  a late eluter)
are recommended.   Deuterated analogs of analytes are not appropriate surrogates
for gas chromatographic/FPD/NPD analysis.

            5.5.1 If surrogates are to be  used, the analyst must select one or
      more compounds that are similar in analytical  behavior to the compounds of
      interest.  The analyst must  further demonstrate that the measurement of a
      surrogate  is  not  affected by method or  matrix  interferences.   General
      guidance on the selection and use of  surrogates is provided  in Sec. 5.0 of
      Method 3500.

            5.5.2 Tributyl  phosphate and  triphenyl phosphate are  used as FPD and
      NPD surrogates.  A  volume of  1.0 mL  of  a  1-^g/L spiking solution (1 ng of
      surrogate)  is added to  each water  sample and each  soil/sediment sample.
      If there is a  co-elution problem,  4-chloro-3-nitrobenzo-trifluoride has
      also been used as an NPD-only surrogate.


6.0   SAMPLE COLLECTION,  PRESERVATION, AND HANDLING

      6.1   See the introductory material  to Chapter Four, "Organic Analytes,"
Sec. 4.0.

      6.2   Extracts are to be refrigerated at  4°C  and analyzed  within 40 days
of extraction.  See Sec.  5.2.4 for storage of standards.

      6.3   Organophosphorus  esters   will   hydrolyze   under  acidic  or  basic
conditions.  Adjust  samples  to a pH of 5 to 8 using  sodium  hydroxide or sulfuric
acid solution as soon as possible after  sample collection.   Record  the volume
used.

      6.4   Even  with storage at  4°C   and  use  of  mercuric  chloride  as  a
preservative,  most  OPs  in  groundwater   samples  collected  for  the  national
pesticide survey  degraded within a 14-day  period.  Begin  sample extraction within
7 days of collection.
                                  8141A - 10                        Revision 1
                                                                September 1994

-------
7.0   PROCEDURE

      7.1   Extraction and cleanup:

            7.1.1 Refer to Chapter Two  and Method 8140 for guidance on choosing
      the appropriate  extraction procedure.   In  general,  water samples  are
      extracted at a  neutral  pH  with methylene chloride,  using  either Method
      3510 or 3520.   Solid  samples  are extracted using either Method  3540 or
      3541 with methylene chloride/acetone (1:1 v/v)  or hexane/acetone (1:1 v/v)
      as the extraction solvent.   Method 3550 is  an  inappropriate  extraction
      technique for the target analytes of this method (See Sec.  2.3).

            7.1.2 Extraction and  cleanup  procedures  that use solutions below pH
      4 or above  pH 8 are  not appropriate for this  method.

            7.1.3 If required, the  samples may be cleaned  up using the  Methods
      presented in Chapter Four,  Sec. 2.  Florisil Column Cleanup (Method 3620)
      and Sulfur  Cleanup (Method  3660,  TBA-sulfite  option)  may have  particular
      application for  OPs.   Gel  Permeation  Cleanup (Method 3640)  should  not
      generally be used for OP pesticides.

                  7.1.3.1      If  sulfur cleanup by  Method  3660 is required, do
            not use mercury or copper.

                  7.1.3.2      GPC may   only   be  employed  if  all  target  OP
            pesticides  are  listed  as   analytes  of  Method  3640,   or  if  the
            laboratory has demonstrated a recovery  of greater  than  85  percent
            for target  OPs  at a concentration not greater  than 5 times  the
            regulatory  action   level.     Laboratories   must   retain   data
            demonstrating  acceptable recovery.

            7.1.4 Prior to gas chromatographic  analysis,  the extraction  solvent
      may be exchanged  to hexane.  The analyst  must ensure quantitative transfer
      of the extract  concentrate.  Single-laboratory data indicate that  samples
      should not  be transferred with  100-percent hexane during sample workup, as
      the more  polar  organophosphorus  compounds may be  lost.   Transfer  of
      organophosphorus esters is  best accomplished  using methylene chloride or
      a hexane/acetone solvent mixture.

            7.1.5 Methylene chloride may be used as an  injection solvent  with
      both the  FPD and the NPD.

            NOTE:  Follow manufacturer's instructions as to suitability of using
                  methylene chloride with any specific detector.

      7.2   Gas chromatographic conditions:

            7.2.1 Four  0.53-mm  ID  capillary  columns are  suggested  for  the
      determination of organophosphates  by this  method.   Column 1  (DB-210  or
      equivalent)  and  Column 2  (SPB-608 or  equivalent)  of  30-m  length  are
      recommended if  a large number of organophosphorus  analytes  are to  be
      determined.   If superior chromatographic  resolution is not required,  15-m
      lengths columns may be appropriate.  Operating  conditions for 15-m  columns


                                 8141A - 11                         Revision 1
                                                               September 1994

-------
      are listed in Table 5.  Operating conditions for 30-m columns are listed
      in Table 6.

            7.2.2 Retention  times  for  analytes on  each  set  of  columns  are
      presented in Tables 3 and 4.

      7.3   Calibration: Refer to Method 8000 for proper calibration techniques.
Use Table 5 and Table 6 for establishing the proper  operating parameters for the
set of columns being employed in the analyses.

      7.4   Gas chromatographic analysis:   Method 8000 provides instructions on
the analysis sequence,  appropriate  dilutions,  establishing  daily retention time
windows and identification criteria.

            7.4.1 Automatic injections  of 1 /uL are recommended.  Hand injections
      of no more  than  2  /it may be used  if the analyst demonstrates quantitation
      precision of < 10  percent  relative standard deviation.  The solvent flush
      technique may be used if the amount of solvent is kept at a minimum.   If
      the internal  standard calibration technique is used,  add 10 /iL of internal
      standard to each mL  of  sample prior  to  injection.   Chromatograms  of  the
      target organophosphorus compounds are shown in Figures 1  through 4.

            7.4.2 Figures 5 and 6 show  chromatograms with and without Simazine,
      Atrazine,  and Carbophenothion on 30-m columns.

      7.5   Record the sample volume  injected  to the nearest 0.05 jiL  and  the
resulting peak sizes  (in  area  units or peak heights).  Using either the internal
or external  calibration  procedure  (Method  8000),  determine the  identity  and
quantity of each  component peak  in  the  sample chromatogram which corresponds to
the compounds used for calibration purposes.   See  Method  8000  for calculation
equations.

            7.5.1 If  peak detection  and  identification   is  prevented by  the
      presence of interferences, the use of an  FPD or further sample cleanup is
      required.   Before using any cleanup procedure, the analyst must process a
      series of calibration standards through the procedure  to establish elution
      patterns and to determine recovery of target  compounds.   The absence of
      interference from reagents must  be demonstrated by  routine processing of
      reagent blanks through  the chosen cleanup  procedure.   Refer to  Sec.  3.0
      for interferences.

            7.5.2 If the  responses  exceed the linear range  of the system,  dilute
      the extract and reanalyze.  It is recommended that extracts be diluted so
      that all peaks  are on scale.  Overlapping peaks are not  always  evident
      when  peaks   are  off-scale.   Computer   reproduction  of  chromatograms,
      manipulated to ensure all peaks  are  on  scale  over  a 100-fold range,  are
      acceptable if linearity is demonstrated.   Peak height measurements  are
      recommended over peak area integration when overlapping peaks cause errors
      in area integration.

            7.5.3 If the  peak response is less than  2.5 times the baseline noise
      level, the validity of the quantitative result may  be questionable.   The
                                  8141A - 12                        Revision 1
                                                                September 1994

-------
      analyst should consult  with  the  source of the sample to determine whether
      further concentration of the sample extract is warranted.

            7.5.4 If partially overlapping  or coeluting peaks are found, change
      columns or try a GC/MS  technique.   Refer to Sec. 8.0 and Method 8270.

      7.6   Suggested chromatograph maintenance: Corrective measures may require
any one or more of the following remedial actions.

            7.6.1 Refer  to  Method   8000  for   general   information  on  the
      maintenance of capillary columns and injectors.

            7.6.2 Splitter connections:   For  dual  columns which  are  connected
      using  a  press-fit Y-shaped glass  splitter or  a  Y-shaped  fused-silica
      connector (J&W Scientific,  Restek, or equivalent),  clean and  deactivate
      the splitter.   Reattach the columns after cleanly cutting  off at least one
      foot from the  injection port side of  the column using a capillary cutting
      tool or  scribe.   The  accumulation of  high boiling residues  can  change
      split ratios between  dual  columns and thereby change calibration factors.

            7.6.3 Columns  will  be  damaged  permanently   and   irreversibly  by
      contact with oxygen at  elevated  temperature.  Oxygen can enter the column
      during a septum change,  when oxygen traps are exhausted, through neoprene
      diaphragms of regulators, and through leaks in  the  gas  manifold.   Polar
      columns  including  the  DB-210 and  DB-608 are more prone  to  oxidation.
      Oxidized  columns  will  exhibit  baselines  that  rise  rapidly  during
      temperature programming.

            7.6.4 Peak tailing for all components will  be exacerbated by dirty
      injectors, pre-columns, and glass  "Y"s.  Additionally,  cleaning  of this
      equipment (or replacement/clipping,  as appropriate) will  greatly  reduce
      the peak  tailing.   Components  such  as  Fensulfothion,  Naled,  Azinphos-
      methyl, and Dimethoate  are very  good indicators  of  system performance.

      7.7   Detector maintenance:

            7.7.1 Older  FPDs  may  be  susceptible   to  stray  light  in  the
      photomultiplier tube compartment.   This stray  light  will decrease  the
      sensitivity and the  linearity of  the detector.   Analysts  can  check  for
      leaks by initiating an  analysis  in  a  dark room and turning on  the lights.
      A  shift  in  the baseline  indicates that light  may  be leaking  into  the
      photomultiplier tube  compartment.   Additional shielding should be applied
      to eliminate light leaks and minimize stray light interference.

            7.7.2 The bead of  the NPD will become exhausted with time,  which
      will decrease the sensitivity and  the selectivity  of  the detector.   The
      collector may become contaminated  which  decreased detector sensitivity.

            7.7.3 Both types  of detectors  use a  flame  to  generate  a  response.
      Flow rates  of air and hydrogen should  be optimized  to give the  most
      sensitive, linear detector response for  target analytes.
                                  8141A -  13                        Revision 1
                                                                September 1994

-------
8.0   QUALITY CONTROL

      8.1   Refer  to Chapter  One  for  specific  quality control  procedures.
Include a mid-level check standard after each group of 10 samples  in the analysis
sequence.  Quality control to validate  sample  extraction is  covered in Method
3500 and in the extraction method utilized.  If extract cleanup was performed,
follow the QC in Method 3600 and in the specific cleanup method.

      8.2   Procedures to check the GC system operation are found  in Method 8000.

      8.3   GC/MS confirmation

            8.3.1 GC/MS techniques  should  be  judiciously employed  to  support
      qualitative  identifications  made  with  this  method.   Follow the  GC/MS
      operating requirements specified in Method 8270.

            8.3.2 When  available,   chemical  ionization  mass spectra  may  be
      employed to aid in the qualitative identification process.

            8.3.3 To confirm an  identification  of  a  compound,  the  background-
      corrected mass spectrum of the compound  must be obtained from the sample
      extract  and  must  be  compared  with  a  mass  spectrum  from  a stock  or
      calibration standard analyzed under the  same chromatographic  conditions.
      At  least  25  ng  of material  should  be  injected   into  the  GC/MS.    The
      following criteria must be met for qualitative confirmation:

                  8.3.3.1     The  qualitative   identification   of   compounds
            determined  by  this  method  is  based  on  retention  time,  and  on
            comparison of the sample mass spectrum,  after  background correction,
            with characteristic   ions   in  a  reference  mass spectrum.    The
            reference mass spectrum must be generated  by the laboratory  using
            the conditions of this  method.   The characteristic  ions from  the
            reference mass spectrum  are defined to be the  three ions of greatest
            relative intensity,  or any ions over 30% relative intensity  if less
            than three  such  ions  occur  in  the reference  spectrum.   Compounds
            should  be identified as present when the criteria below are met.

                        8.3.3.1.1   The  intensities of  the characteristic ions
                  of a compound  maximize in the  same scan or within one  scan of
                  each  other.   Selection of  a peak by  a data  system  target
                  compound search  routine  where  the  search  is  based  on  the
                  presence of  a target  chromatographic  peak containing  ions
                  specific  for   the  target  compound  at a  compound-specific
                  retention  time will be accepted as meeting  this criterion.

                        8.3.3.1.2   The  RRT of  the  sample component is  within
                  ± 0.06 RRT units of the RRT  of the standard component.

                        8.3.3.1.3   The     relative     intensities    of    the
                  characteristic  ions  agree  within  30%   of   the relative
                  intensities  of  these   ions   in   the   reference   spectrum.
                  (Example:    For  an ion  with  an abundance  of 50%  in  the
                  reference  spectrum, the corresponding  abundance in a  sample
                  spectrum can range between  20% and 80%.)

                                  8141A  -  14                         Revision  1
                                                                September 1994

-------
            8.3.3.1.4   Structural isomers that produce very similar
      mass  spectra  should  be identified  as  individual  isomers if
      they  have   sufficiently  different   GC   retention  times.
      Sufficient  GC  resolution is  achieved  if the  height  of the
      valley between two isomer peaks  is less than 25% of the sum of
      the  two  peak  heights.    Otherwise, structural  isomers are
      identified as isomeric pairs.

            8.3.3.1.5   Identification  is  hampered  when  sample
      components are not resolved  chromatographically and produce
      mass  spectra  containing  ions  contributed  by  more  than one
      analyte.   When gas chromatographic  peaks obviously represent
      more  than one  sample  component  (i.e.,  a broadened peak with
      shoulder(s)  or  a  valley  between  two  or  more  maxima),
      appropriate  selection  of  analyte   spectra  and  background
      spectra is  important.   Examination  of  extracted ion current
      profiles  of  appropriate  ions  can  aid  in  the  selection of
      spectra,  and in qualitative identification of compounds.  When
      analytes  coelute  (i.e.,  only  one  chromatographic peak is
      apparent), the identification criteria  can  be  met,  but each
      analyte spectrum will  contain extraneous ions contributed by
      the coeluting compound.

      8.3.3.2     For samples containing components not associated
with the calibration  standards,  a library search may be made for the
purpose of tentative identification.   The  necessity to perform this
type of  identification  will be determined by the purpose  of the
analyses  being  conducted.     Computer  generated  library  search
routines   should   not   use   normalization   routines   that   would
misrepresent the library or  unknown  spectra  when  compared to each
other.   For example, the RCRA permit or waste delisting requirements
may require the reporting of nontarget analytes.  Only after visual
comparison of sample spectra with the  nearest library searches will
the  mass   spectral  interpretation  specialist  assign a  tentative
identification. Guidelines for making  tentative identification are:

      (1)   Relative intensities  of  major  ions  in  the  reference
spectrum (ions  > 10% of  the  most abundant  ion) should be present in
the sample  spectrum.

      (2)   The relative intensities  of  the major ions should agree
within + 20%.   (Example:  For an ion with an abundance of 50% in the
standard spectrum, the  corresponding  sample  ion  abundance must be
between 30 and 70%.)

      (3)   Molecular ions present  in  the  reference spectrum should
be present  in the sample spectrum.

      (4)   Ions present  in the  sample  spectrum  but not  in  the
reference  spectrum  should  be  reviewed  for  possible  background
contamination or presence of coeluting compounds.

      (5)   Ions present in the reference spectrum but  not  in the
sample spectrum should be reviewed for possible subtraction from the

                      8141A  - 15                        Revision 1
                                                    September 1994

-------
            sample  spectrum  because of background  contamination  or coeluting
            peaks.  Data system library reduction  programs can sometimes create
            these discrepancies.

            8.3.4 Where  available,  chemical  ionization mass  spectra  may be
      employed to aid in the qualitative identification process because of the
      extensive  fragmentation  of organophosphorus  pesticides  during electron
      impact MS processes.

            8.3.5 Should the MS procedure  fail  to  provide satisfactory results,
      additional steps may be taken  before reanalysis.   These steps may include
      the use of alternate packed or capillary GC columns or additional sample
      cleanup.


9.0   METHOD PERFORMANCE

      9.1   Estimated MDLs and associated chromatographic conditions for water
and clean soil (uncontaminated with synthetic organics) are listed in Table 1.
As detection limits  will vary with the particular matrix to be analyzed, guidance
for determining EQLs is given in Table 2.  Recoveries for several method analytes
are provided in Tables 5, 6, and 7.


10.0  REFERENCES

1.    Taylor, V.; Hickey, D.M.; Marsden, P.J. "Single Laboratory Validation of
      EPA Method  8140"; U.S.  Environmental  Protection  Agency,  Environmental
      Monitoring Systems  Laboratory, Office of Research and  Development,  Las
      Vegas, NV, 1987; EPA-600/4-87-009.

2.    Pressley, T.A;  Longbottom,  J.E.   "The  Determination  of  Organophosphorus
      Pesticides  in Industrial  and Municipal  Wastewater:  Method  614";  U.S.
      Environmental   Protection Agency, Environmental  Monitoring  and  Support
      Laboratory, Cincinnati, OH, 1982; EPA-600/4-82-004.

3.    "Analysis of  Volatile  Hazardous  Substances by  GC/MS:  Pesticide  Methods
      Evaluation";  Letter  Reports  6,  12A,  and  14  to  the  U.S.  Environmental
      Protection Agency on Contract 68-03-2697, 1982.

4.    "Method 622, Organophosphorus Pesticides";  U.S. Environmental Protection
      Agency, Environmental  Monitoring  and  Support  Laboratory,  Cincinnati, OH
      45268.

5.    Lopez-Avila,  V.;  Baldin,  E.;  Benedicto,  J; Milanes,  J.;  Beckert,  W. F.
      "Application of Open-Tubular Columns to SW-846 GC Methods"; final report
      to the U.S. Environmental Protection Agency on Contract 68-03-3511; Mid-
      Pacific Environmental  Laboratory, Mountain  View, CA,  1990.

6.    Hatcher,  M.D.;    Hickey, D.M.;   Marsden,   P.O.;  and   Betowski,  L.D.;
      "Development of a GC/MS Module for RCRA Method 8141"; final report to the
      U.S. EPA Environmental  Protection Agency  on  Contract 68-03-1958; S-Cubed,
      San Diego, CA, 1988.


                                  8141A -  16                        Revision 1
                                                                September 1994

-------
7.    Chau, A.S.Y.; Afghan,  B.K.  Analysis of Pesticides  in Water; "Chlorine and
      Phosphorus-Containing Pesticides"; CRC:  Boca Raton, FL, 1982, Vol. 2, pp
      91-113, 238.

8.    Hild,  J.;  Schulte,   E;  Thier,   H.P.   "Separation  of  Organophosphorus
      Pesticides   and   Their    Metabolites  on   Glass-Capillary   Columns";
      Chromatographia, 1978, 11-17.

9.    Luke,  M.A.;  Froberg,  J.E.;  Doose,  G.M.;  Masumoto,  H.T.  "Improved
      Multiresidue  Gas  Chromatographic  Determination  of  Organophosphorus,
      Organonitrogen,   and  Organohalogen  Pesticides   in  Produce,  Using  Flame
      Photometric and  Electrolytic Conductivity Detectors"; J. Assoc, Off. Anal.
      Chem. 1981, 1187, 64.

10.   Sherma, J.;  Berzoa,   M.  "Analysis  of Pesticide  Residues  in Human  and
      Environmental Samples";  U.S.  Environmental  Protection Agency,  Research
   •   Triangle Park, NC; EPA-600/8-80-038.

11.   Desmarchelier,  J.M.;   Wustner,   D.A.;  Fukuto,  T.R.  "Mass  Spectra  of
      Organophosphorus Esters and Their Alteration Products"; Residue Reviews,
      1974, pp 63, 77.

12.   Munch, D.J. and   Frebis, C.P., "Analyte Stability Studies Conducted during
      the National Pesticide Survey",  ES & T,  1992,  vol  26, 921-925.

13.   T.L. Jones, "Organophosphorus Pesticide Standards:  Stability Study", EMSL-
      LV Research Report, EPA 600/X-92/040, April, 1992

14.   Kotronarou, A.,  et al., "Decomposition of Parathion in Aqueous Solution by
      Ultrasonic Irradiation,"  ES&T,  1992, Vol.  26,  1460-1462.
                                  8141A - 17                        Revision 1
                                                                September 1994

-------
                              TABLE 1
           METHOD  DETECTION  LIMITS IN  A WATER AND A SOIL
    MATRIX USING 15-m COLUMNS AND A FLAME PHOTOMETRIC DETECTOR
Compound
                              Reagent
                              Water (3510)"
Soil (3540)
Azinphos-methyl
Bolstar (Sulprofos)
Chlorpyrifos
Coumaphos
Demeton, -0, -S
Diazinon
Dichlorvos (DDVP)
Dimethoate
Disulfoton
EPN
Ethoprop
Fensulfothion
Fenthion
Malathion
Merphos
Mevinphos
Naled
Parathion, ethyl
Parathion, methyl
Phorate
Ronnel
Sulfotepp
TEPPC
Tetrachlorovinphos
Tokuthion (Protothiofos)0
Trichloronatec
0.10
0.07
0.07
0.20
0.12
0.20
0.80
0.26
0.07
0.04
0.20
0.08
0.08
0.11
0.20
0.50
0.50
0.06
0.12
0.04
0.07
0.07
0.80
0.80
0.07
0.80
5.0
3.5
5.0
10.0
6.0
10.0
40.0
13.0
3.5
2.0
10.0
4.0
5.0
5.5
10.0
25.0
25.0
3.0
6.0
2.0
3.5
3.5
40.0
40.0
5.5
40.0
Sample  extracted  using  Method  3510,   Separatory  Funnel  Liquid-Liquid
Extraction.

Sample extracted using Method 3540, Soxhlet Extraction.

Purity  of  these  standards  not  established  by  the  EPA  Pesticides  and
Industrial  Chemicals Repository, Research Triangle Park, NC.
                            8141A - 18
        Revision 1
    September 1994

-------
                                    TABLE  2
                DETERMINATION  OF  ESTIMATED QUANTITATION  LIMITS
                         (EQLs) FOR VARIOUS MATRICES8
   Matrix                                                            Factor
   Ground water (Methods 3510 or 3520)                                 10
   Low-concentration soil by Soxhlet and no cleanup                    10
   Non-water miscible waste (Method 3580)                            1000
a  EQL = [Method detection limit (see  Table  1)] X  [Factor found in this table].
For non-aqueous samples, the factor is on a wet-weight basis.  Sample EQLs are
highly matrix dependent.  The EQLs  to  be  determined herein are for guidance and
may not always be achievable.

b  Multiply this factor times the reagent water MDL in Table 1.

c  Multiply this factor times the soil MDL in Table 1.
                                  8141A - 19                        Revision 1
                                                                September 1994

-------
                TABLE  3.
RETENTION TIMES FOR METHOD 8141A ANALYTES
          EMPLOYING  15-m COLUMNS

TEPP
Dichlorvos (DDVP)
Mevinphos
Demeton, -0 and -S
Ethoprop
Naled
Phorate
Monochrotophos
Sulfotepp
Dimethoate
Disulfoton
Diazinon
Merphos
Ronnel
Chlorpyrifos
Malathion
Parathion, methyl
Parathion, ethyl
Trichloronate
Tetrachlorovinphos
Tokuthion (Protothiofos)
Fensulfothion
Bolstari (Sulprofos)
Famphur"
EPN
Azinphos-methyl
Fenthion
Coumaphos
Method 8141A has not been fully
Initial temperature
Initial time
Program 1 rate
Program 1 final temp.
Program 1 hold
Program 2 rate
Program 2 final temp.
Program 2 hold
Capi
Compound

9.63
14.18
18.31
18.62
19.94
20.04
20.11
20.64
23.71
24.27
26.82
29.23
31.17
31.72
31.84
31.85
32.19
34.65
34.67
35.85
36.34
36.40
38.34
38.83
39.83
llary Column
DB-5
6.44
7.91
12.88
15.90
16.48
19.01
17.52
20.11
18.02
20.18
19.96
20.02
21.73
22.98
26.88
28,78
23.71
27.62
28.41
32.99
24.58
35.20
35.08
36.93
37.80
38.04
29.45
38.87

SPB-608
5.12
12.79
18.44
17.24
18.67
17.40
18.19
31.42
19.58
27.96
20.66
19.68
32.44
23.19
25.18
32.58
32.17
33.39
29.95
33.68
39.91
36.80
37.55
37.86
36.71
37.24
28.86
39.47
DB-210
10.66
19.35








36.74


val idated- for Famphur.
130°C
3 minutes
5°C/min
180°C
10 minutes
2°C/min
250°C
15 minutes
50°C
1 minute
5°C/min
140°C
10 minutes
10°C/min
240°C
10 minutes
50°C
1 minute
5°C/min
140°C
10 minutes
10°C/min
240°C
10 minutes


            8141A  -  20
    Revision 1
September 1994

-------
                                   TABLE  4.
                   RETENTION  TIMES  FOR  METHOD 8141A ANALYTES
                            EMPLOYING 30-m COLUMNS8
Compound
DB-5
      RT (min)

DB-210      DB-608
DB-1
Trimethyl phosphate
Dichlorvos (DDVP)
Hexamethylphosphoramide
Trichlorfon
TEPP
Thionazin
Mevinphos
Ethoprop
Diazinon
Sulfotepp
Terbufos
Tri-o-cresyl phosphate
Naled
Phorate
Fonophos
Disulfoton
Merphos
Oxidized Merphos
Dichlorofenthion
Chlorpyrifos, methyl
Ronnel
Chlorpyrifos
Trichloronate
Aspon
Fenthion
Demeton-S
Demeton-0
Monocrotophosc
Dimethoate
Tokuthion
Malathion
Parathion, methyl
Fenithrothion
Chlorfenvinphos
Parathion, ethyl
Bolstar
Stirophos
Ethion
b
7.45
b
11.22
b
12.32
12.20
12.57
13.23
13.39
13.69
13.69
14.18
12.27
14.44
14.74
14.89
20.25
15.55
15.94
16.30
17.06
17.29
17.29
17.87
11.10
15.57
19.08
18.11
19.29
19.83
20.15
20 .£3
21.07
21.38
22.09
22.06
22.55
2.36
6.99
7.97
11.63
13.82
24.71
10.82
15.29
18.60
16.32
18.23
18.23
15.85
16.57
18.38
18.84
23.22
24.87
20.09
20.45
21.01
22.22
22.73
21.98
22.11
14.86
17.21
15.98
17.21
24.77
21.75
20.45
21.42
23.66
22.22
27.57
24.63
27.12

6.56

12.69


11.85
18.69
24.03
20.04
22.97

18.92
20.12

23.89

35.16
26.11
26.29
27.33
29.48
30.44

29.14
21.40
17.70
19.62
20.59
33.30
28.87
25.98

32.05
29.29
38.10
33.40
37.61

10.43




14.45
18.52
21.87
19.60


18.78
19.65

21.73
26.23



23.67
24.85


24.63
20.18

19.3
19.87
27.63
24.57
22.97


24.82
29.53
26.90

                                                                   (continued)
                                  8141A - 21
                            Revision 1
                        September 1994

-------
                             TABLE 4. (Continued)
Compound
 DB-5
       RT (min)

 DB-210      DB-608      DB-1
Phosphamidon
Crotoxyphos
Leptophos
Fensulfothion
EPN
Phosmet
Azinphos-methyl
Azinphos-ethyl
Famphur
Coumaphos
Atrazine
Simazine
Carbophenothion
Dioxathion
Trithion methyl
Dicrotophos
Internal Standard
l-Bromo-2-nitrobenzene
Surrogates
Tributyl phosphate
Triphenyl phosphate
4-C1-3-nitrobenzotrifluoride
22.77
22.77
24.62
27.54
27.58
27.89
28.70
29.27
29.41
33.22
13.98
13.85
22.14
 d
 8.11
5.73
20.09
23.85
31.32
26.76
29.99
29.89
31.25
32.36
27.79
33.64
17.63
17.41
27.92
 9.07
 5.40
25.88
32.65
44.32
36.58
41.94
41.24
43.33
45.55
38.24
48.02
              22.24
              36.62
              19.33
                       11.1
                       33.4
28.58
31.60

32.33
34.82
a The GC operating conditions were as follows:

DB-5 and DB-210 - 30-m  x  0.53-mm  ID  column,  DB-5 (1.50-  m film thickness) and
DB-210 (1.0- m film thickness).  Both connected to a  press-fit Y-shaped inlet
splitter.   Temperature  program:  120°C  (3-min hold) to 270"C  (10-min  hold) at
5°C/min; injector  temperature 250°C; detector  temperature 300°C; bead temperature
400"C; bias voltage  4.0;  hydrogen gas pressure  20 psi;  helium carrier  gas  6
mL/min; helium makeup gas 20 mL/min.

DB-608 - 30-m x 0.53-mm ID column,  DB-608  (1.50-  m  film thickness) installed in
an 0.25-in  packed-column  inlet.   Temperature program:  110°C (0.5-min  hold) to
250°C (4-min hold) at 3°C/min;  injector temperature 250°C; helium carrier gas 5
mL/min; flame photometric detector.

DB-1 30-m x 0.32-mm ID column, DB-1  (0.25-  m film thickness) split/splitless with
head pressure of  10  psi,  split valve closure at  45 sec,  injector  temp.  250°C,
50°C (1-min hold)  to 280°C (2-min  hold) at 6°C/min, mass spectrometer full scan
35-550 amu.

b Not detected at 20 ng per injection.
c Retention times  may shift to longer times with larger amounts  injected (shifts
of over 30 seconds have been observed,  Hatcher et. al.)
d Shows multiple peaks; therefore, not included  in the composite.
                                  8141A - 22
                              Revision  1
                          September  1994

-------
                               TABLE  5.
PERCENT RECOVERY OF 27 ORGANOPHOSPHATES BY SEPARATORY FUNNEL EXTRACTION
Compound
Azinphos methyl
Bolstar
Chlorpyrifos
Coumaphos
Demeton
Diazinon
Dichlorvos
Dimethoate
Disulfoton
EPN
Ethoprop
Fensulfonthion
Fenthion
Malathion
Merphos
Mevinphos
Monocrotophos
Naled
Parathion, ethyl
Parathion, methyl
Phorate
Ronnel
Sulfotep
TEPP
Tetrachlorvinphos
Tokuthion
Trichloroate

Low
126
134
7
103
33
136
80
NR
48
113
82
84
NR
127
NR
NR
NR
NR
101
NR
94
67
87
96
79
NR
NR
Percent Recovery
Medium
143 + 8
141 + 8
89 + 6
90 + 6
67 + 11
121 + 9.5
79 + 11
47 + 3
92 + 7
125 + 9
90 + 6
82 + 12
48 + 10
92 + 6
79
NR
18 + 4
NR
94 + 5
46 + 4
77 + 6
97 + 5
85 + 4
55 + 72
90 + 7
45 + 3
35

High
101
101
86
96
74
82
72
101
84
97
80
96
89
86
81
55
NR
NR
86
44
73
87
83
63
80
90
94
NR = Not recovered.
                            8141A - 23
    Revision 1
September 1994

-------
                                   TABLE 6.
PERCENT RECOVERY OF 27 ORGANOPHOSPHATES BY CONTINUOUS LIQUID-LIQUID EXTRACTION
Percent Recovery
Compound
Azinphos methyl
Bolstar
Chlorpyrifos
Coumaphos
Demeton
Diazinon
Dichlorvos
Dimethoate
Disulfoton
EPN
Ethoprop
Famphur
Fensulfonthion
Fenthion
Malathion
Merphos
Mevinphos
Monocrotophos
Naled
Parathion, ethyl
Parathion, methyl
Phorate
Ronnel
Sulfotep
TEPP
Tetrachlorvinphos
Tokuthion
Trichloroate
Low
NR
NR
13
94
38
NR
81
NR
94
NR
39
._
90
8
105
NR
NR
NR
NR
106
NR
84
82
40
39
56
132
NR
Medium
129
126
82 + 4
79 + 1
23 + 3
128 + 37
32 + 1
10 + 8
69 + 5
104 + 18
76 + 2
63 + 15
67 + 26
32 + 2
87 + 4
80
87
30
NR
81 + 1
50 + 30
63 + 3
83 + 7
77 + 1
18 + 7
70 + 14
32 + 14
NR
High
122
128
88
89
41
118
74
102
81
119
83
--
90
86
86
79
49
1
74
87
43
74
89
85
70
83
90
21
    NR = Not recovered.
                                8141A -  24
    Revision 1
September 1994

-------
                               TABLE 7.
     PERCENT  RECOVERY  OF  27  ORGANOPHOSPHATES  BY  SOXHLET EXTRACTION
Percent Recovery
Compound
Azinphos methyl
Bolstar
Chlorpyrifos
Coumaphos
Demeton
Diazinon
Dichlorvos
Dimethoate
Disulfoton
EPN
Ethoprop
Fensulfonthion
Fenthion
Malathion
Merphos
Mevinphos
Monocrotophos
Naled
Parathion, ethyl
Parathion, methyl
Phorate
Ronnel
Sulfotep
TEPP
Tetrachlorvinphos
Tokuthion
Trichloroate
Low
156
102
NR
93
169
87
84
NR
78
114
65
72
NR
100
62
NR
NR
NR
75
NR
75
NR
67
36
50
NR
56
Medium
110 + 6
103 + 15
66 + 17
89 + 11
64 + 6
96 + 3
39 + 21
48+7
78 + 6
93 + 8
70 + 7
81 + 18
43 + 7
81 + 8
53
71
NR
48
80 + 8
41 + 3
77 + 6
83 + 12
72 + 8
34 + 33
81 + 7
40 + 6
53
High
87
79
79
90
75
75
71
98
76
82
75
111
89
81
60
63
NR
NR
80
28
78
79
78
63
83
89
53
NR = Not recovered.
                            8141A - 25
    Revision 1
September 1994

-------
                               TABLE 8.

            SUGGESTED OPERATING CONDITIONS FOR 15-m COLUMNS
Columns 1 and 2 (DB-210 and SPB-608 or their equivalent)
Carrier gas (He) flow rate
Initial temperature =
Temperature program =
Column 3  (DB-5 or equivalent)

Carrier gas (He) flow rate =
Initial temperature =
Temperature program =
5 mL/min
50°C, hold for 1 minute
50°C to 140°C at 5°C/min, hold for
10 minutes,  followed by  140°C  to
240°C  at  10°C/min,   hold  for  10
minutes (or a sufficient amount of
time for last compound to elute).
5 mL/min
130°C, hold for 3 minutes
130°C to 180°C at  5°C/min, hold for
10 minutes,  followed by  180°C  to
250°C  at  2°C/min,  hold  for  15
minutes (or a sufficient amount of
time for last compound to elute).
                             8141A  -  26
                    Revision 1
                September 1994

-------
                                      TABLE 9
                  SUGGESTED  OPERATING  CONDITIONS FOR 30-m COLUMNS
Column 1:
      Type:   DB-210
      Dimensions:   30-m x 0.53-mm ID
      Film Thickness  (jum):  1.0

Column 2:

      Type:   DB-5
      Dimensions:   30-m x 0.53-mm ID
      Film Thickness  (p.m):  1.5


Carrier gas flowrate (mL/min):  6 (Helium)

Makeup gas flowrate (mL/min):  20 (Helium)

Temperature program:   120°C (3-min hold) to 270°C (10-min hold) at 5°C/min

Injector temperature:  250°C

Detector temperature:  300°C

Injection volume:   2 /^L

Solvent:  Hexane

Type of injector:   Flash vaporization

Detector type:  Dual  NPD

Range:  1

Attenuation:  64

Type of splitter:    Y-shaped  or Tee

Data system:   Integrator

Hydrogen gas pressure:  20 psi

Bead temperature:   400°C

Bias voltage:  4
                                  8141A - 27                        Revision 1
                                                                September 1994

-------
                               TABLE 10
        QUANTITATION AND CHARACTERISTIC IONS FOR OP PESTICIDES
Compound Name
Quantitation ions
Characteristic ions
Azinphos-methyl
Bolstar (Sulprofos)
Chlorpyrifos
Coumaphos
Demeton-S
Diazinon
Dichlorvos (DDVP)
Dimethoate
Disulfoton
EPN
Ethoprop
Fensulfothion
Fenthion
Malathion
Merphos
Mevinphos
Monocrotophos
Naled
Parathion, ethyl
Parathion, methyl
Phorate
Ronnel
Stirophos
Sulfotepp
TEPP
Tokuthion
160
156
197
109
88
137
109
87
88
157
158
293
278
173
209
127
127
109
291
109
75
285
109
322
99
113
77,132
140,143,113,33
97,199,125.314
97,226,362,21
60,114,170
179,152,93,199,304
79,185,145
93,125,58,143
89,60,61,97,142
169,141,63,185
43,97,41,126
97,125,141,109,308
125,109,93,169
125,127,93,158
57,153,41,298
109,67,192
67,97,192,109
145,147,79
97,109,139,155
125,263,79
121,97,47,260
125,287,79,109
329,331,79
97,65,93,121,202
155,127,81,109
43,162,267,309
                            8141A - 28
                                          Revision 1
                                      September 1994

-------
300.X
250.00
200.00-
150.00
100.00
 50.00
  0.00
I

faratmnn eim/i
Tetrachlorovinphos
Fensulfothion
• ' V... Ji


                   A  „.
         11.. I ... 11. 111 » I •. 111 • I,.,.,,,. ,1. • I 11, (,.. p ,,,.,, I ,., I,,.,,., I ...,. , ,, ... I ..,,.,.,.,,,
         1   3   5   7   9 11  13  15  17 19  21 23  25 27  29 31  33  35  37  39  41  43  45
 Figure 1.   Chromatogram of  target organophosphorus  compounds  from a  15-m DB-210
 column with  FPD detector.   More compounds are  shown  in Figure  2.  See  Table 3 for
 retention times.
                                    8141A  - 29
    Revision  1
September  1994

-------
300.00
250.00
200.00
150.00
100.00
 50.00
  0.00
                                o
                                i
                                o
                                   \
                                                          (A
                                                          O

                                                          Q.
                                                          O
                                                          O
                                                            I
0.
lil
>.
I
                                                                         Q.
         1   3  5   7  9  11  13  15  17 19  21  23 25  27  29 31  33  35 37  39 41 43  45
 Figure  2.   Chromatogram  of  target  organophosphorus  compounds  from  a 15-m  DB-210
 column  with FPD detector.   More compounds  are shown  in  Figure 1.   See Table 3 for
 retention times.
                                     8141A - 30
    Revision  1
September  1994

-------
300.00
250.00
200.00 H
150.00 H
100.00 H
 50.00 H
  0.00-
t-p-iTT rrrrt-rrriTrptT|i t ij t . i; n T p > »rTTT t1T T I" ' I '" I ' " I "TT T" I" ' I"M ' T'
 1   3  5   7  9  11 13  15   17  19 21  23 25  27  29 31  33 35 37  39 41  43  45
  Figure 3.   Chromatogram  of target  organophosphorus  compounds  from a  15-m DB-210
  column with  NPD detector.   More compounds  are  shown in  Figure 4. See  Table 3 tor
  retention times.
                                     8141A  -  31
                                                                 Revision 1
                                                             September 1994

-------
300.00 -,
250.00-
200.00-
150.00-
100.00-
 50.00-
  0.00-
        1   3   5  7  9   11  13  15 17  19  21  23  25 27  29 31  33  35  37  39 41  43  45
  Figure 4.   Chromatogram  of target  organophosphorus  compounds  from  a 15-m  DB-210
  column with NPD detector.  More compounds  are shown  in  Figure 3.   See Table  3  for
  retention times.
                                     8141A - 32
    Revision  1
September 1994

-------
      -f
                                                             DB-210
                                                           OB-5
Figure 5.  Chromatogram of target organophosphorus compounds on a 30-m DB-5/DB-210
column pair with NPD detector,  without Simazine, Atrazine and Carbophenothion.  See
Table 4 for retention times and for GC operating conditions.
                                  8141A - 33
    Revision 1
September 1994

-------
     -r
                                                  DB-210
                                        It
                                    It
                                    It
                                          r-fcvN
         I         <

        ^•« f—JWtAi
                                                            DB-5

Figure 6.  Chromatogram of target organophosphorus compounds on  a 30-m DB-5/DB-210
column pair with  NPD  detector,  with  Simazine,  Atrazine and  Carbophenothion.   See
Table 4 for retention times and for GC operating conditions.
                                  8141A - 34
    Revision 1
September 1994

-------
                                       METHOD  8141A
               ORGANOPHOSPHORUS  COMPOUNDS  BY  GAS  CHROMATOGRAPHY:
                              CAPILLARY  COLUMN  TECHNIQUE
        Start
         I
7.1.1 Refer to Chapter
  Two for guidance on
choosing the appropriate
  extraction procedure.
     7.1 .2 Perform
   solvent exchange
      during K-D
   procedures in all
  extraction methods.
        I
     7.2 Select GC
      conditions.
  7.3 Refer to Method
       8000 for
 calibration techniques.
   7.3.1  Internal or
       external
    calibration  may
       be used.
        I
  7.4.1 Add internal
  standard to sample
     if necessary.
     7.4.2 Refer to
   Method 8OOO, Sec.
   7.6 for instructions
  on analysis sequence,
dilutions, retention times,
    and  identification
        criteria.
  7.4.3 Inject sample.
  7.4.5 Record sample
  volume injected and
  resulting peak sizes.
    7.4.6 Determine
      identity and
    quantity of each
    component peak;
    refer to Method
   8000, Sec. 7.8 for
  calculation equations.
        7.4.7
        Is paak
     detection and
     identification
     prevented b'y
       interfer
        encos?
   7.5.1 Perform
appropriate cleanup.
                                  7.5.2 Reanalyze by
                                         GC.
  c
                                        1
                                                                        Stop
                                     8141A -  35
                                                 Revision  1
                                            September  1994

-------
00
I-*
I/I
o

-------
                                 METHOD 8150A

                 CHLORINATED HERBICIDES BY GAS CHROMATOGRAPHY
1.0   SCOPE AND APPLICATION

      1.1   Method 8150  is  a  gas chromatographic  (GC)  method  for determining
certain chlorinated acid herbicides.  The  following compounds can be determined
by this method:
      Compound Name                          CAS No.a


      2,4-0                                  94-75-7
      2,4-DB                                 94-82-6
      2,4,5-TP (Silvex)                      93-72-1
      2,4,5-T                                93-76-5
      Dalapon                                75-99-0
      Dicamba                              1918-00-9
      Dichlorprop                           120-36-5
      Dinoseb                                88-85-7
      MCPA                                   94-74-6
      MCPP                                   93-65-2


      a  Chemical  Abstract  Services  Registry  Number.

     1.2   Table  1  lists  the  method  detection  limit  for  each compound  in
organic-free reagent  water.  Table 2  lists the estimated quantitation limit (EQL)
for other matrices.

     1.3   When  Method 8150  is  used  to analyze  unfamiliar  samples,  compound
identifications  should  be  supported  by at  least one  additional  qualitative
technique.   This  method describes  analytical   conditions  for  a second  gas
chromatographic column that can  be  used to confirm  measurements  made  with the
primary column.  Section 8.4 provides gas chromatograph/mass spectrometer (GC/MS)
criteria   appropriate   for    the  qualitative   confirmation   of   compound
identifications.

     1.4   Only experienced analysts should be allowed to work with diazomethane
due to the potential  hazards  associated with  its  use  (the compound is explosive
and carcinogenic).


2.0  SUMMARY OF METHOD

     2.1   Method   8150   provides   extraction,   esterification,   and   gas
chromatographic  conditions  for the analysis  of  chlorinated  acid herbicides.
Spiked  samples are used  to  verify the  applicability  of the  chosen  extraction
technique to each new sample type.   The esters  are  hydrolyzed with potassium
hydroxide, and extraneous organic material  is  removed by a solvent wash.  After

                                  8150A -  1                         Revision 1
                                                                     July 1992

-------
acidification, the  acids are extracted with  solvent and converted to their methyl
esters using diazomethane as  the  derivatizing  agent.   After excess  reagent is
removed,  the esters are determined by gas chromatography employing an electron
capture  detector,    microcoulometric detector,  or electrolytic  conductivity
detector (Goerlitz  and Lamar,  1967).   The results  are  reported as  the  acid
equivalents.

     2.2    The  sensitivity of  Method 8150  usually depends  on the  level  of
interferences rather than on instrumental limitations.
3.0   INTERFERENCES

      3.1    Refer  to  Method 8000.

      3.2    Organic acids,  especially  chlorinated  acids,  cause the most direct
interference with  the determination.  Phenols,  including chlorophenols, may also
interfere with this procedure.

      3.3    Alkaline  hydrolysis  and subsequent extraction of the basic solution
remove many chlorinated hydrocarbons and phthalate esters that might otherwise
interfere with the electron capture analysis.

      3.4    The  herbicides,  being  .strong  organic acids,  react  readily  with
alkaline substances and may be lost during analysis.   Therefore, glassware and
glass wool  must  be  acid  rinsed,  and  sodium sulfate  must be  acidified  with
sulfuric acid prior to use to avoid this possibility.


4.0   APPARATUS AND MATERIALS

      4.1    Gas  chromatograph

            4.1.1  Gas  chromatograph,   analytical   system  complete  with  gas
      chromatograph   suitable   for  on-column   injections   and   all   required
      accessories, including detectors, analytical  columns,  recorder, gases, and
      syringes.  A data system for measuring  peak  heights and/or peak areas is
      recommended.

            4.1.2  Columns

                  4.1.2.1      Column  la and Ib  -  1.8 m x 4 mm ID glass, packed
            with  1.5%  SP-2250/1.95% SP-2401  on  Supelcoport  (100/120  mesh) or
            equivalent.

                  4.1.2.2      Column 2 - 1.8 m  x  4 mm ID glass, packed with 5%
            OV-210 on Gas  Chrom  Q (100/120  mesh) or equivalent.

                  4.1.2.3      Column  3  -  1.98  m  x 2  mm  ID glass, packed  with
            0.1% SP-1000 on 80/100 mesh  Carbopack  C or equivalent.

            4.1.3  Detector - Electron  capture (ECD).

      4.2    Erlenmeyer flasks - 250 and 500 mL  Pyrex,  with 24/40 ground glass

                                   8150A -  2                         Revision  1
                                                                     July  1992

-------
joint.

     4.3   Beaker  - 500 ml.

     4.4   Diazomethane generator  - Refer  to  Section 7.4 to  determine  which
method of diazomethane generation should be used for a particular application.

           4.4.1  Diazald  kit  -  recommended for the generation of diazomethane
     using the procedure given in Section  7.4.2  (Aldrich Chemical Co., Cat. No.
     210,025-2 or equivalent).

           4.4.2  Assemble from  two  20 x 150 mm test tubes, two Neoprene rubber
     stoppers, and  a  source of  nitrogen.   Use Neoprene  rubber  stoppers  with
     holes drilled in them to accommodate glass delivery tubes.  The exit tube
     must be drawn to  a point to bubble  diazomethane through the sample extract.
     The generator assembly is shown in Figure 1.  The  procedure for use of this
     type of generator is given  in Section 7.4.3.

     4.5   Vials  -  10 to  15 ml, amber  glass,  with Teflon lined  screw  cap  or
crimp top.

     4.6   Separatory funnel  -  2000 ml, 125 ml,  and 60 ml.

     4.7   Drying  column  - 400 mm x 20 mm ID Pyrex chromatographic column with
Pyrex glass wool  at bottom and a Teflon stopcock.

     NOTE: Fritted  glass  discs  are difficult  to decontaminate  after  highly
           contaminated extracts have been  passed through.    Columns  without
           frits  may  be  purchased.   Use  a small  pad of Pyrex  glass  wool  to
           retain  the adsorbent.   Prewash the glass wool  pad with 50  ml  of
           acetone  followed  by  50  ml of elution  solvent  prior to packing the
           column  with  adsorbent.

     4.8   Kuderna-Danish (K-D)  apparatus

           4.8.1  Concentrator tube  - 10 ml, graduated (Kontes K-570050-1025 or
     equivalent).   A  ground glass  stopper  is  used to prevent evaporation  of
     extracts

           4.8.2  Evaporation  flask - 500 ml (Kontes K-570001-500 or equivalent).
     Attach to concentrator tube with springs, clamps or  equivalent.

           4.8.3  Snyder  column  -  Three ball  macro  (Kontes  K-503000-0121  or
     equivalent).

           4.8.4  Snyder  column  -   Two ball  micro  (Kontes  K-569001-0219  or
     equivalent).

           4.8.5  Springs  -   1/2 inch  (Kontes  K-662750 or equivalent).

     4.9   Boiling chips  - Solvent  extracted,  approximately 10/40 mesh (silicon
carbide or equivalent).

     4.10  Water   bath  -  Heated,   with  concentric  ring cover,  capable  of

                                   8150A  -  3                        Revision  1
                                                                     July 1992

-------
temperature control  (* 5°C).   The bath should be used in a hood.
     4.11  Microsyringe  -  10  ^L.
     4.12  Wrist  shaker  -  Burrell Model  75  or  equivalent.
     4.13  Glass  wool  -  Pyrex,  acid washed.
     4.14  Balance  - Analytical, capable of  accurately weighing to the nearest
0.0001 g.
     4.15  Syringe  - 5 ml.
     4.16  Glass  rod.

5.0  REAGENTS
     5.1   Reagent  grade chemicals shall  be  used in all tests.  Unless otherwise
indicated, it is intended that all reagents  shall conform  to the specifications
of the Committee on  Analytical Reagents of the  American Chemical Society, where
such specifications are  available.   Other grades may  be  used,  provided it is
first ascertained that the reagent is  of  sufficiently high purity to permit its
use without lessening the accuracy of the determination.
     5.2   Organic-free  reagent water.   All  references to  water in this method
refer to organic-free reagent water, as defined in Chapter One.
     5.3   Sulfuric acid solution
           5.3.1  ((1:1)  (v/v))  - Slowly  add  50 ml H2S04 (sp. gr. 1.84) to 50 ml
     of organic-free reagent  water.
           5.3.2  ((1:3)  (v/v))  - Slowly  add  25 ml H2S04 (sp. gr. 1.84) to 75 ml
     of organic-free reagent  water.
     5.4   Hydrochloric  acid  ((1:9) (v/v)),HCl.  Add one volume of concentrated
HC1 to 9 volumes  of organic-free reagent water.
     5.5   Potassium  hydroxide  solution  (KOH)  -  37%  aqueous  solution (w/v).
Dissolve 37  g  potassium hydroxide pellets  in  organic-free reagent water, and
dilute to 100 ml.
     5.6   Carbitol (Diethylene glycol monoethyl  ether),  C2H5OCH2CH2OCH2CH2OH.
Available from Aldrich Chemical Co.
     5.7   Solvents
           5.7.1  Acetone, CH3COCH3  -  Pesticide quality or equivalent.
           5.7.2  Methanol, CH3OH -  Pesticide quality or equivalent.
           5.7.3  Hexane, C6HU  - Pesticide quality or equivalent.

                                   8150A  -  4                        Revision  1
                                                                     July  1992

-------
            5.7.4  Diethyl  Ether,  C2H5OC2H5.   Pesticide quality  or equivalent.
     Must  be  free  of  peroxides as  indicated  by  test  strips  (EM  Quant,  or
     equivalent).   Procedures  for  removal of peroxides  are  provided with the
     test strips.   After cleanup,  20 ml  of ethyl  alcohol preservative must be
     added to each  liter of ether.

     5.8    Sodium  sulfate  (granular,  anhydrous),  Na2SO,.  Purify by heating at
400°C  for 4  hours in a  shallow tray,  or by precleaning the sodium sulfate with
methylene chloride.  If the  sodium sulfate  is precleaned with methylene chloride,
a method blank must be analyzed, demonstrating that there is no interference from
the sodium sulfate.

     5.9    N-Methyl-N-nitroso-p-toluenesulfonamide(Diazald),CH3C6H4S02N(CH3)NO.
 Available from Aldrich Chemical  Co.

     5.10   Silicic  acid.   Chromatographic grade,  nominal 100 mesh.   Store at
130°C.

     5.11   Stock standard  solutions - Stock standard solutions can be prepared
from pure standard materials or purchased as certified solutions.

            5.11.1       Prepare stock standard solutions  by accurately weighing
     about 0.0100 g of pure acids.   Dissolve the material in pesticide quality
     diethyl ether  and  dilute to  volume  in a 10 ml  volumetric  flask.  Larger
     volumes can be used at the convenience of the analyst.  If compound purity
     is certified at 96% or greater,  the weight can be used without correction
     to  calculate  the  concentration  of  the  stock  standard.    Commercially
     prepared  stock standards can be  used at  any  concentration  if they are
     certified by the manufacturer or by  an independent  source.

            5.11.2       Transfer the stock  standard  solutions  into vials with
     Teflon lined screw caps  or  crimp tops.   Store at  4°C and  protect from
     light.  Stock standard solutions  should be  checked frequently for signs of
     degradation or evaporation,  especially just prior to preparing calibration
     standards from them.

            5.11.3       Stock standard  solutions must be  replaced  after 1 year,
     or  sooner if comparison with check standards indicates a problem.

     5.12   Calibration  standards - A minimum of five  calibration standards  for
each parameter of interest should be  prepared  through  dilution  of the stock
standards  with  diethyl  ether.   One  of  the  concentrations should be  at   a
concentration  near, but above,  the method detection  limit.    The  remaining
concentrations should correspond to the expected range of concentrations found
in  real  samples  or should  define  the working  range of  the GC.   Calibration
solutions must be replaced  after six months,  or  sooner  if comparison  with check
standards indicates a problem.

     5.13   Internal standards (if  internal standard  calibration  is  used) - To
use this approach,  the analyst must select one or more internal standards that
are similar in analytical  behavior to  the compounds  of  interest.  The analyst
must further demonstrate that  the  measurement  of  the internal  standard  is not
affected by method  or  matrix  interferences.   Because of  these limitations, no
internal standard can be suggested that is applicable to all samples.

                                   8150A - 5                         Revision  1
                                                                      July  1992

-------
            5.13.1       Prepare  calibration  standards  at  a  minimum of  five
     concentrations for each parameter of interest as described in Section 5.12.

            5.13.2       To  each  calibration  standard,  add  a  known  constant
     amount of one or more internal standards, and dilute to volume with diethyl
     ether.

            5.13.3       Analyze each calibration standard according to Section
     7.0.

     5.14   Surrogate standards - The analyst should monitor the performance of
the extraction,  cleanup (when used), and analytical  system and the effectiveness
of  the  method  in  dealing  with  each sample  matrix  by  spiking each  sample,
standard,  and  organic-free  reagent water  blank  with  one  or two  herbicide
surrogates (e.g. herbicides  that  are  not expected to be present in the sample).
The surrogates selected should elute over the range of the temperature program
used in  this  method.    Deuterated  analogs  of analytes  should not be used  as
surrogates for gas chromatographic analysis due  to coelution  problems.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1    See  the introductory material  to this Chapter,  Organic  Analytes,
Section 4.1.  Extracts  must  be stored under  refrigeration and  analyzed within 40
days of extraction.


7.0  PROCEDURE

     7.1    Preparation of waste  samples

            7.1.1  Extraction

                  7.1.1.1      Follow Method 3580 except use diethyl  ether as the
            dilution solvent, acidified anhydrous sodium sulfate, and acidified
            glass  wool.

                  7.1.1.2      Transfer 1.0  mL (a lesser volume or a dilution may
            be required  if herbicide concentrations are high)  to a 250 mL ground
            glass-stoppered   Erlenmeyer   flask.     Proceed   to  Section   7.2.2
            hydrolysis.

     7.2    Preparation of soil,  sediment,  and other  solid  samples

            7.2.1  Extraction

                  7.2.1.1      To a 500 mL,  wide mouth Erlenmeyer  flask  add 50
            g  (dry weight) of the well mixed,  moist solid sample.  Adjust the pH
            to  2 with  concentrated HC1 and monitor  the  pH  for 15  minutes with
            occasional  stirring.   If  necessary,  add additional  HC1  until  the pH
            remains at  2.

                  7.2.1.2      Add  20 mL  acetone  to  the  flask  and mix  the
            contents with  the wrist shaker for  20  minutes.   Add 80 mL diethyl

                                   8150A  -  6                         Revision  1
                                                                     July 1992

-------
      ether to the same flask and shake again for 20 minutes.   Decant  the
      extract and measure the volume of solvent recovered.

            7.2.1.3      Extract  the  sample twice  more  using  20 mL  of
      acetone followed by 80 ml of diethyl  ether.   After addition of each
      solvent, the  mixture  should  be  shaken with  the  wrist  shaker  for
      10 minutes and the acetone-ether extract decanted.

            7.2.1.4      After the  third extraction, the volume of extract
      recovered should be at least  75%  of the volume of added solvent.   If
      this  is not  the case,  additional  extractions may be  necessary.
      Combine the extracts in a 2  liter  separatory funnel  containing 250  ml
      of 5% acidified sodium sulfate.   If an emulsion  forms,  slowly add 5
      g of  acidified  sodium sulfate (anhydrous) until  the  solvent-water
      mixture separates.  A  quantity of acidified  sodium sulfate equal  to
      the weight of the sample may be added, if necessary.

            7.2.1.5      Check the pH  of the extract.   If it  is not at  or
      below pH 2,  add more concentrated HC1  until stabilized at the desired
      pH.  Gently mix  the contents  of  the  separatory  funnel  for 1 minute
      and allow the  layers  to  separate.   Collect the aqueous  phase in a
      clean beaker  and  the  extract phase  (top  layer)  in  a 500  ml ground
      glass-stoppered Erlenmeyer flask. Place the aqueous phase back into
      the separatory funnel  and  re-extract  using 25 ml  of diethyl ether.
      Allow the layers to separate  and  discard the aqueous layer.  Combine
      the ether extracts in  the 500 ml Erlenmeyer flask.

      7.2.2 Hydrolysis

            7.2.2.1      Add 30 mL of  organic-free  reagent water, 5 ml  of
      37% KOH, and  one  or two  clean boiling  chips  to  the flask.   Place a
      three ball Snyder column on  the  flask, evaporate  the diethyl ether
      on a water bath,  and continue to heat for a total  of 90 minutes.

            7.2.2.2     Remove the flask from the water bath and allow to
      cool.  Transfer the water solution to  a 125 ml separatory funnel  and
      extract the basic solutions once with 40 mL and then  twice with 20  ml
      of diethyl ether.  Allow sufficient time  for the layers to separate
      and discard the ether  layer each time. The  phenoxy-acid herbicides
      remain  soluble in the aqueous phase as potassium  salts.

      7.2.3 Solvent cleanup

      7.2.3.1      Adjust  the pH  to 2  by adding 5  mL  cold (4°C)  sulfuric
acid (1:3)  to the  separatory  funnel.   Be  sure to  check the pH  at this
point.    Extract  the herbicides once with  40 mL  and  twice with  20 mL  of
diethyl  ether.  Discard the  aqueous phase.

            7.2.3.2     Combine ether  extracts in a 125 mL  Erlenmeyer
      flask  containing  5-7 g  of  acidified anhydrous   sodium  sulfate.
      Stopper and allow the  extract to remain in  contact with the acidified
      sodium  sulfate.   If concentration and esterification are not to  be
      performed   immediately,    store   the  sample   overnight   in   the
      refrigerator.

                             8150A - 7                         Revision 1
                                                                July  1992

-------
            NOTE:  The  drying  step is very critical to  ensuring  complete
                   esterification.   Any moisture  remaining  in the  ether
                   will  result  in  low herbicide  recoveries.  The amount  of
                   sodium sulfate is adequate  if  some free flowing  crystals
                   are  visible when swirling the flask.   If all  the  sodium
                   sulfate solidifies in a  cake,  add a few additional  grams
                   of acidified sodium sulfate and again test by swirling.
                   The  2 hour  drying  time is  a minimum,  however, the
                   extracts  may be  held  overnight  in   contact  with the
                   sodium sulfate.

            7.2.3.3      Transfer  the  ether   extract,  through  a  funnel
      plugged with acid washed glass wool,  into a 500 ml  K-D flask  equipped
      with a  10  ml concentrator tube.   Use  a  glass rod  to crush  caked
      sodium sulfate during  the transfer.   Rinse the Erlenmeyer flask and
      column with 20-30 ml of diethyl ether to  complete the  quantitative
      transfer.

            7.2.3.4      Add  one  or two clean boiling chips to the  flask
      and attach a three ball  Snyder column.   Prewet the Snyder column  by
      adding about 1 mL of  diethyl  ether to the  top.  Place  the apparatus
      on  a  hot water  bath  (60°-65°C)  so  that  the  concentrator  tube  is
      partially immersed in  the  hot water and  the entire lower rounded
      surface  of  the  flask  is bathed  in vapor.   Adjust  the  vertical
      position of the apparatus and the water temperature,  as required,  to
      complete the concentration  in 15-20  minutes.   At  the proper rate  of
      distillation the  balls  of the column will  actively chatter, but the
      chambers will not flood.  When the apparent volume of liquid reaches
      1 ml,  remove the  K-D apparatus from the water bath  and allow  it  to
      drain  and cool for at least  10 minutes.

            7.2.3.5      Remove the Snyder column  and rinse  the flask and
      its lower joints  into the concentrator tube with  1-2 mL of diethyl
      ether.   A  5 ml  syringe is recommended  for  this  operation.   Add  a
      fresh  boiling chip, attach a micro Snyder column to the concentrator
      tube,  and prewet  the column by adding 0.5 ml of  ethyl  ether  to the
      top.  Place  the  micro K-D apparatus on  the water bath so  that the
      concentrator tube is partially immersed in  the hot water.  Adjust the
      vertical position  of the apparatus and  the  water  temperature  as
      required  to complete  concentration in  5-10 minutes.    When the
      apparent volume of the liquid reaches 0.5 mL, remove  the micro K-D
      from the bath  and allow it  to drain and  cool.   Remove the  Snyder
      column  and  add  0.1  mL of  methanol.    Rinse   the walls  of the
      concentrator tube while adjusting the extract volume to 1.0 mL with
      diethyl ether.  Proceed to Section 7.4 for esterification.

7.3   Preparation of aqueous samples

      7.3.1  Extraction

            7.3.1.1      Using  a  1  liter graduated  cylinder,  measure  1
      liter (nominal) of sample,  record the sample volume to  the nearest
      5   mL,   and   transfer   it   to  the  separatory   funnel.   If  high
      concentrations are anticipated, a smaller volume may  be used and then

                             8150A - 8                         Revision 1
                                                               July 1992

-------
diluted with organic-free reagent water to  1  liter.  Adjust  the  pH
to less than 2 with sulfuric acid (1:1).

      7.3.1.2      Add  150 ml of diethyl ether to the sample bottle,
seal, and  shake  for  30 seconds to  rinse  the walls.  Transfer  the
solvent wash  to  the separatory  funnel  and  extract the sample  by
shaking the funnel for  2  minutes  with periodic venting to  release
excess pressure.   Allow  the  organic layer to separate from  the water
layer for  a minimum of 10 minutes.  If the emulsion interface between
layers is  more than  one third the  size of the solvent layer,  the
analyst must  employ  mechanical  techniques to  complete the  phase
separation.  The optimum technique depends upon the  sample  and may
include stirring,  filtration  of the  emulsion through  glass  wool,
centrifugation,  or other physical  methods.   Drain  the aqueous phase
into a 1 liter Erlenmeyer flask.  Collect the  solvent extract in  a
250 ml ground glass  Erlenmeyer flask containing  2  ml  of 37% KOH.
Approximately 80  ml of the diethyl  ether will remain dissolved in the
aqueous phase.

      7.3.1.3      Repeat  the  extraction two more  times  using 50  ml
of diethyl ether each time.   Combine the extracts  in the Erlenmeyer
flask.   (Rinse  the 1 liter flask with  each  additional aliquot  of
extracting solvent.)

7.3.2 Hydrolysis

      7.3.2.1      Add  one or two clean boiling chips and  15 ml  of
organic-free reagent water  to  the  250 ml  flask and  attach  a three
ball Snyder column.  Prewet the Snyder  column by  adding about 1  ml
of diethyl ether to the top of the  column.   Place the  apparatus  on
a hot water bath  (60°-65°C) so  that the bottom  of the flask is bathed
with hot water vapor.  Although the  diethyl  ether will  evaporate in
about  15   minutes,  continue heating  for   a  total  of  60  minutes,
beginning  from  the time  the  flask  is  placed  in the  water bath.
Remove the apparatus and let stand at room temperature  for at least
10 minutes.

      7.3.2.2      Transfer the solution to a 60 ml separatory funnel
using 5-10 ml of organic-free reagent water.  Wash  the basic solution
twice by shaking for 1  minute  with 20 ml portions  of diethyl  ether.
Discard the  organic  phase.   The  herbicides  remain   in  the  aqueous
phase.

7.3.3 Solvent cleanup

      7.3.3.1      Acidify the contents  of the separatory  funnel  to
pH 2 by adding 2  ml of  cold  (4°C) sulfuric acid (1:3).  Test with pH
indicator  paper.  Add 20 ml diethyl  ether  and shake vigorously  for
2 minutes.  Drain the aqueous layer into a 250 ml Erlenmeyer flask,
and pour the organic  layer into a 125 ml Erlenmeyer flask containing
about 5-7 g of acidified sodium sulfate.  Repeat the extraction twice
more with  10 ml aliquots of diethyl  ether,  combining all solvent in
the  125 ml flask.  Allow  the  extract  to remain in contact with  the
sodium sulfate for approximately 2 hours.

                       8150A - 9                         Revision 1
                                                          July 1992

-------
             NOTE:  The drying step is very  critical  to  ensuring  complete
                   esterification.   Any  moisture  remaining in the  ether
                   will  result  in  low herbicide recoveries.  The amount of
                   sodium sulfate is adequate if some  free flowing  crystals
                   are visible  when swirling the flask.  If all the sodium
                   sulfate solidifies in a cake, add a few additional  grams
                   of acidified sodium sulfate  and again test by swirling.
                   The  2  hour   drying  time  is  a minimum,  however,  the
                   extracts may  be  held  overnight   in  contact with  the
                   sodium sulfate.

             7.3.3.2      Transfer the  ether  extract,   through  a  funnel
      plugged with acid washed  glass wool, into a 500  ml K-D flask  equipped
      with  a 10 ml  concentrator  tube.   Use a glass rod to  crush  caked
      sodium sulfate during the transfer.  Rinse  the  Erlenmeyer flask and
      column with  20-30 ml of  diethyl ether to complete the quantitative
      transfer.

             7.3.3.3      Add  one or two clean boiling chips to the  flask
      and attach a three ball Snyder  column.  Prewet  the Snyder column by
      adding about 1 ml of diethyl  ether to the top.   Place the apparatus
      on  a  hot water  bath (60°-65°C)  so that the  concentrator   tube  is
      partially  immersed  in  the hot water  and the   entire lower  rounded
      surface  of  the  flask  is  bathed  in  vapor.    Adjust the  vertical
      position of the apparatus and the water temperature, as required, to
      complete the concentration in  15-20 minutes.  At the proper rate of
      distillation the balls  of the  column will actively chatter,  but the
      chambers will not flood.   When the apparent volume of liquid reaches
      1 ml,  remove the  K-D apparatus from the  water  bath  and  allow  it to
      drain  and cool for at least 10 minutes.

             7.3.3.4      Remove the  Snyder column  and rinse the flask and
      its lower joints  into  the concentrator tube with  1-2 ml of  diethyl
      ether.   A 5 mL syringe  is  recommended  for  this operation.   Add a
      fresh boiling chip, attach a micro Snyder column to the concentrator
      tube,  and prewet  the column by adding 0.5 ml of ethyl ether to the
      top.   Place  the  micro  K-D apparatus on the water bath  so  that the
      concentrator tube is partially immersed in the hot  water.  Adjust the
      vertical  position of  the  apparatus   and the water temperature  as
      required  to complete  concentration  in  5-10  minutes.    When  the
      apparent volume of  the liquid reaches 0.5 ml,  remove  the micro K-D
      from  the  bath  and allow it to drain  and cool.   Remove the Snyder
      column  and  add   0.1  ml  of methanol.   Rinse the  walls  of  the
      concentrator tube while adjusting  the extract volume to  1.0 ml with
      diethyl ether.

7.4   Esterification

      7.4.1  Two  methods may  be used  for the generation  of  diazomethane:
the bubbler method (set up shown in  Figure 1)  and  the Diazald  kit  method.
The bubbler method  is  suggested  when  small batches (10-15)  of  samples
require esterification.  The bubbler method works well  with samples that
have low concentrations of herbicides  (e.g.  aqueous samples) and  is  safer
to use than the Diazald  kit procedure.  The  Diazald kit method  is  good for

                             8150A -  10                         Revision 1
                                                               July 1992

-------
are
large quantities of samples needing esterification.   The Oiazald kit method
is more effective than  the  bubbler method for soils or  samples  that may
contain high concentrations  of  herbicides  (e.g. samples such as soils that
may result  in  yellow extracts following  hydrolysis may be  difficult  to
handle by the bubbler method).  The diazomethane derivatization (U.S. EPA,
1971) procedures, described below, will react efficiently with all of the
chlorinated herbicides  described  in this method and  should be used only by
experienced analysts, due to the potential  hazards associated with  its use.
The following precautions should be taken:

      CAUTION:    Diazomethane is  a  carcinogen  and can  explode  under
                  certain   conditions.

             Use a safety screen.
             Use mechanical  pipetting  aides.
             Do  not heat above  90°C --  EXPLOSION may result.
             Avoid grinding surfaces, ground glass joints, sleeve bearings,
             glass stirrers  --     EXPLOSION may result.
             Store away  from alkali metals  --  EXPLOSION may result.
             Solutions of diazomethane decompose rapidly in the presence of
             solid materials such  as  copper powder,  calcium chloride, and
             boiling  chips.

      7.4.2  Diazald  kit method -  Instructions for  preparing  diazomethane
    provided with the generator kit.

             7.4.2.1      Add 2 ml of diazomethane solution and let sample
      stand  for 10 minutes with occasional swirling.

             7.4.2.2      Rinse inside wall of ampule with several hundred
      pi  of diethyl  ether.   Allow solvent to evaporate  spontaneously at
      room  temperature  to about 2  mL.

             7.4.2.3      Dissolve the residue in  5  mL of hexane.   Analyze
      by  gas chromatography.

      7.4.3  Bubbler  method  -  Assemble  the diazomethane  bubbler  (see
Figure 1).

             7.4.3.1      Add 5 mL of diethyl  ether to the first test tube.
      Add 1  mL  of diethyl ether,  1 mL of carbitol,   1.5 mL of 37% KOH, and
      0.1-0.2 g Diazald to the second test  tube.   Immediately place the
      exit tube into the concentrator tube containing the sample extract.

             Apply nitrogen  flow (10 mL/min) to bubble diazomethane through
      the extract for 10 minutes or until  the yellow color of diazomethane
      persists.    The  amount   of   Diazald  used  is   sufficient  for
      esterification of approximately three sample extracts. An additional
      0.1-0.2  g of  Diazald may  be  added  (after the initial  Diazald is
      consumed) to  extend  the generation  of  the diazomethane.   There is
      sufficient KOH present in the original  solution to perform a maximum
      of  approximately  20 minutes  of  total esterification.

             7.4.3.2      Remove the  concentrator tube and  seal  it with  a

                             8150A - 11                        Revision  1
                                                                July 1992

-------
           Neoprene or Teflon stopper.  Store at room temperature  in a hood for
           20 minutes.

                  7.4.3.3      Destroy  any  unreacted  diazomethane  by  adding
           0.1-0.2 g  silicic  acid to  the  concentrator tube.   Allow  to stand
           until  the evolution of nitrogen gas  has stopped.  Adjust the sample
           volume  to  10.0 mL  with hexane.   Stopper  the  concentrator  tube and
           store  refrigerated if  further  processing will  not  be  performed
           immediately.    It  is  recommended  that  the methylated extracts  be
           analyzed immediately to minimize the trans-esterification and other
           potential reactions that may occur.  Analyze by gas chromatography.

     7.5   Gas chromatographic conditions  (Recommended)

           7.5.1  Column  la

           Carrier gas (5% methane/95% argon) flow rate:     70  mL/min
           Temperature program:     185°C,  isothermal.

           7.5.2  Column  Ib
           Carrier gas  (5% methane/95% argon) flow rate:    70 mL/min
           Initial temperature:     140°C, hold for 6 minutes
           Temperature  program:     140°C to 200°C at  10°C/min, hold until last
                                    compound has eluted.

           7.5.3  Column 2

           Carrier gas  (5% methane/95% argon) flow rate:    70 mL/min
           Temperature  program:     185°C, isothermal.

                  7.5.4  Column 3

           Carrier gas  (ultra-high purity N2) flow rate:    25 mL/min
           Initial temperature:     100°C, no hold
           Temperature  program:     100°C to 150°C at  10°C/min, hold until last
                                    compound has eluted.

     7.6   Calibration  - Refer to Method 8000 for proper calibration techniques.
Use Table 1 and especially Table 2 for guidance on selecting  the lowest point on
the calibration curve.

           7.6.1  The procedure for internal or external calibration may be used.
     Refer to Method 8000 for a description of each of these procedures.

           7.6.2  The following  gas chromatographic columns  are recommended for
     the compounds indicated:

           Analyte
           Dicamba
           2,4-D
           2,4,5-TP
           2,4,5-T
           2,4-DB

                                  8150A - 12                        Revision 1
                                                                     July 1992

-------
            Daiapon                              3
            MCPP                                 lb
            MCPA                                 lb
            Dichloroprop                         lb
            Oinoseb                              lb

     7.7    Gas chromatographic  analysis

            7.7.1  Refer to Method 8000.   If  the  internal  standard calibration
     technique is used, add  10  pi  of internal  standard to the sample prior to
     injection.

            7.7.2  Method  8000 provides instructions on  the  analysis  sequence,
     appropriate  dilutions,  establishing  daily   retention  time  windows,  and
     identification criteria.  Include a mid-concentration check standard after
     each group of 10  samples in the analysis sequence.

            7.7.3  Examples of chromatograms for various chloro-phenoxy herbicides
     are shown in Figures 2 through 4.

            7.7.4  Record the  sample volume injected and  the resulting peak sizes
     (in area units or peak heights).

            7.7.5  Using either the internal  or external  calibration  procedure
     (Method 8000),  determine the identity and quantity of each component peak-
     in the sample  chromatogram which corresponds  to  the compounds  used  for
     calibration purposes.

            7.7.6  If calibration standards have been analyzed  in the same manner
     as the  samples (e.g. have  undergone  hydrolysis and esterification), then
     the calculation  of  concentration given in  Method  8000 should  be  used.
     However,  if  calibration is done  using  standards  made  from  methyl  ester
     compounds (compounds not esterified  by  application  of  this method), then
     the  calculation   of  concentration  must  include   a  correction  for  the
     molecular weight  of  the methyl ester versus the acid herbicide.

            7.7.7  If  peak detection  and   identification  are  prevented due  to
     interferences,  further  cleanup  is  required.  Before  using  any  cleanup
     procedure, the  analyst must  process a series of standards  through  the
     procedure to validate  elution patterns and  the absence of interferences
     from reagents.


8.0  QUALITY CONTROL

     8.1    Refer  to  Chapter One  for  specific   quality  control  procedures.
Quality control to validate sample extraction is  covered in Method 3500 and in
the extraction method  utilized.  If extract cleanup was  performed,  follow the QC
in Method 3600 and in  the specific cleanup method.

     8.2    Procedures  to check the  GC  system operation  are found  in Method 8000,
Section 8.6.

            8.2.1  Select  a representative  spike concentration for each compound

                                  8150A - 13                        Revision 1
                                                                     July 1992

-------
     (acid or ester) to be measured.  Using stock standards, prepare a quality
     control check sample concentrate in acetone 1,000 times more concentrated
     than the selected concentrations.

           8.2.2  Table  3  indicates  Single  Operator Accuracy and Precision for
     this method.   Compare  the results obtained  with  the results  given  in
     Table 3 to determine if the data quality is acceptable.

     8.3   Calculate  surrogate  standard recovery on all  samples,  blanks,  and
spikes.   Determine  if  the  recovery is within  limits  (limits  established  by
performing QC procedures outlined in Method 8000, Section 8.10).

           8.3.1  If recovery is not within  limits,  the following procedures are
     required.

            •      Check  to  be  sure  there  are no  errors  in  calculations,
                  surrogate  solutions  and   internal  standards.   Also,  check
                  instrument performance.

            •      Recalculate  the data and/or reanalyze  the  extract  if any  of
                  the  above  checks  reveal a problem.

            •      Reextract  and  reanalyze the sample  if none of the above are a
                  problem  or flag the data  as  "estimated  concentration".

     8.4    GC/MS  confirmation

           8.4.1  GC/MS  techniques  should  be  judiciously employed  to support
     qualitative  identifications made with this method.   Refer to Method 8270
     for the appropriate GC/MS operating conditions and  analysis procedures.

            8.4.2  When  available,   chemical  ionization  mass  spectra may  be
     employed to  aid the qualitative identification process.

            8.4.3  Should  these  MS  procedures  fail  to  provide  satisfactory
     results, additional steps may be taken before  reanalysis.  These steps may
     include the  use of alternate packed or capillary GC columns or additional
     cleanup.


9.0  METHOD PERFORMANCE

     9.1    In  a  single  laboratory,  using organic-free   reagent  water  and
effluents from publicly owned treatment works  (POTW),  the  average recoveries
presented in  Table  3  were obtained.   The  standard deviations  of the percent
recoveries of these measurements are also  included in Table 3.
10.0 REFERENCES

1.   U.S. EPA, National  Pollutant Discharge Elimination System,  Appendix A, Fed.
     Reg., 38, No.  75,  Pt.  II,  Method for Chlorinated Phenoxy Acid Herbicides
     in  Industrial  Effluents, Cincinnati, Ohio, 1971.


                                  8150A - 14                        Revision 1
                                                                     July 1992

-------
     Goerlitz, D.G., and W.L. Lamar, "Determination of Phenoxy Acid Herbicides
     in Water by Electron  Capture  and Microcoulometric  Gas  Chromatography,"
     U.S.  Geol. Survey Water Supply Paper, 1817-C, 1967.

3.    Burke, J.A.,  "Gas Chromatography for  Pesticide  Residue Analysis;   Some
     Practical Aspects,"   Journal  of  the Association of  Official  Analytical
     Chemists, 48, 1037, 1965.

4.    U.S.   EPA, "Extraction  and Cleanup  Procedure for  the Determination  of
     Phenoxy Acid Herbicides in  Sediment," EPA Toxicant and  Analysis Center, Bay
     St. Louis, Mississippi, 1972.

5.    "Pesticide Methods Evaluation," Letter Report  #33  for  EPA Contract No. 68-
     03-2697. Available from U.S. Environmental  Protection Agency, Environmental
     Monitoring and Support Laboratory, Cincinnati, Ohio 45268.

6.    Eichelberger, J.W.,  L.E.  Harris,  and W.L. Budde,  "Reference  Compound to
     Calibrate   Ion   Abundance  Measurement   in   Gas   Chromatography-Mass
     Spectrometry," Analytical  Chemistry, 47, 995, 1975.

7.    Glaser,  J.A.  et.al.,   "Trace  Analysis  for  Wastewaters,"  Environmental
     Science  & Technology, 15,  1426, 1981.

8.    U.S.   EPA, "Method 615.  The Determination  of Chlorinated  Herbicides in
     Industrial and Municipal Wastewater," Environmental Monitoring and Support
     Laboratory, Cincinnati, Ohio,  45268, June 1982.
                                  8150A - 15                        Revision 1
                                                                     July 1992

-------
                                   TABLE 1.
                CHROMATOGRAPHIC  CONDITIONS  AND  DETECTION  LIMITS
                           FOR CHLORINATED  HERBICIDES
Retention

Compound
2,4-D
2,4-DB
2,4,5-T
2,4,5-TP (Silvex)
Dalapon
Dicamba
Dichloroprop
Dinoseb
MCPA
MCPP

Col. la
2.0
4.1
3.4
2.7
-
1.2
-
-
-
-

Col.lb
.
-
-
-
-
-
4.8
11.2
4.1
3.4
time (rnin)8

Col. 2 Col. 3
1.6
-
2.4
2.0
5.0
1.0
-
-
-
- -
Method
detection
1 imit (M9/U
1.2
0.91
0.20
0.17
5.8
0.27
0.65
0.07
249
192
  Column conditions are given in Sections 4.1  and 7.5.
                                   TABLE 2.
                    DETERMINATION  OF  ESTIMATED  QUANTITATION
                      LIMITS (EQL) FOR VARIOUS MATRICES8
    Matrix
  Factor
Ground water (based on one liter sample size)
Soil/sediment and other solids
Waste samples
     10
    200
100,000
"Sample EQLs are highly matrix dependent. The EQLs listed herein are provided for
guidance and may not always be achievable.

bEQL - [Method detection limit (Table  1)]  X [Factor  (Table 2)]. For non-aqueous
samples, the factor is  on a wet weight basis.
                                  8150A - 16
        Revision 1
         July 1992

-------
                                   TABLE 3.
                    SINGLE  OPERATOR ACCURACY AND PRECISION8
Compound
2,4-D


Dalapon


2,4-DB


Dicamba


Dichlorprop


Dinoseb

MCPA


MCPP


2,4,5-T


2,4,5-TP


Sample
Type
DW
MM
MW
OW
MW
MW
DW
MW
MW
DW
MW
MW
DW
MW
MW
MW
MW
DW
MW
MW
DW
MW
MW
DW
MW
MW
DW
MW
MW
Spike
(M9/D
10.9
10.1
200
23.4
23.4
468
10.3
10.4
208
1.2
1.1
22.2
10.7
10.7
213
0.5
102
2020
2020
21400
2080
2100
20440
1.1
1.3
25.5
1.0
1.3
25.0
Mean
Recovery
(%)
75
77
65
66
96
81
93
93
77
79
86
82
97
72
100
86
81
98
73
97
94
97
95
85
83
78
88
88
72
Standard
deviation
W
4
4
5
8
13
9
3
3
6
7
9
6
2
3
2
4
3
4
3
2
4
3
2
6
4
5
5
4
5
"All  results based upon seven replicate analyses.  Esterification performed using
the bubbler method. Data obtained from reference 9.

DW = ASTM Type II
MW - Municipal water
                                  8150A - 17
Revision 1
 July 1992

-------
                                FIGURE 1.
                          DIAZOMETHANE GENERATOR
                                                                 gloss lubing
    nitrogen
rubber  stopper
                     tube 1
lubo 2
                                  8150A -  18
                        Revision  1
                         July  1992

-------
                    FIGURE  2.
  GAS  CHROMATOGRAM OF  CHLORINATED HERBICIDES
Column: 1.5% 9-2250/1.95% SP-2401 on Soptlcopon (100/120 M«*>
Ttmp«rttur«: ItottMrmal it 18S°C
D«ttctor: Electron C*ptjr«
          012345
            HCTENTION TIME (MINUTES)
                    8150A  -  19
Revision  1
 July  1992

-------
                  FIGURE  3.
GAS CHROMATOGRAM  OF CHLORINATED HERBICIDES
     Column: 1.5% SP-2250/1.95% SP-2401 on Suptlcoport (100/120 M«*)
     Program: 140°C for 6 Min. 10°C/Minuti to 200°C
     Ottcctor: Electron Capturt
           468
           RETENTION TIME (MINUTES)
10
12
                   8150A - 20
            Revision 1
             July  1992

-------
                    FIGURE 4.
     GAS CHROMATOGRAM  OF DALAPON, COLUMN  3
   I
          Column: 0.1% 9-1000 en 10/100 M«h Ortxxnk C
                 100°C, 10°C/Min to 150°C
                : Electron Captur*
                I
   0246
ftfTfMTION T1MC (MINUTES)
                    8150A - 21
Revision 1
 July 1992

-------
                              METHOD 8150A
          CHLORINATED HERBICIDES BY GAS  CHROMATOGRAPHY
    7211
 Adjust  sample
 pH with HC1
7212 Extract
 sample three
  times with
  acetone and
diethyl ether
    7214
    Combine
   extracts
 7215 Check
pH of  extract.
   adjust if
  necessary.
Separate layers
                                                   Mas te
                                                   sample
   7215
 Re-extract
 and  discard
   aqueous
   phase
  7  1  1 1 Folio.
  Method 3580 for
 extraction, using
  diethyl ether.
acidified anhydrous
sodium sulfate and
  acidified glass
      wool
7 2  2  Proceed
    with
 hydro lysis
    ?  1  L 2 Use
     V 0 ml of
    sampl• for
    hydrolysis
7 2  3 Proceed
«ith 101v«n t
  cleanup
                                                7 3  1 Extract
                                                 three times
                                                • ith dielhyi
                                                   olh.r
   7313
   Combine
  ax tracts
7  3  2 Proceed
    with
 hydr o1ys u
                           733 Proceed
                           with solvent
                             cleanup
                                  8150A  -  22
                                                         Revision  1
                                                           July 1992

-------
                                METHOD  8150A
                                 (Continued)
?43  Axsembe
diazomethane
  oubb1«r.
  genera ie
dla zome thane
                           7  5 Sal
                       chroma tograohic
                         cond 111on»
                        7  6  Clatbra*. e
                        acco rdi ng '.o
                        Mathod 8000
                        7 6  2  Choose
                         appropriate
                         CC  column
742 Preoara
d la z om« lhan*
acco rding to
    ' tanda rds
                      through syj t*m
                         c I aanup
                                   8150A  -  23
                              Revision  1
                                July 1992

-------
00
o
w

-------
                                 METHOD 8150B

                 CHLORINATED HERBICIDES BY GAS CHROMATOGRAPHY
1.0   SCOPE AND APPLICATION

      1.1   Method 8150  is  a  gas chromatographic  (GC) method  for determining
certain chlorinated acid herbicides.  The following compounds can be determined
by this method:
      Compound Name                         CAS No.'
      2,4-D                                  94-75-7
      2,4-DB                                 94-82-6
      2,4,5-TP (Silvex)                      93-72-1
      2,4,5-T                                93-76-5
      Dalapon                                75-99-0
      Dicamba                              1918-00-9
      Dichloroprop                          120-36-5
      Dinoseb                                88-85-7
      MCPA                                   94-74-6
      MCPP                                   93-65-2
      a  Chemical  Abstract  Services  Registry  Number.

      1.2   Table  1  lists  the  method  detection  limit  for  each compound  in
organic-free reagent water.  Table 2  lists the estimated quantitation limit (EQL)
for other matrices.

      1.3   When  Method  8150  is used to analyze  unfamiliar  samples,  compound
identifications should  be  supported  by at  least one  additional  qualitative
technique.   This  method  describes  analytical  conditions  for  a second  gas
chromatographic column that can  be  used to confirm  measurements  made  with  the
primary column.  Sec.  8.4 provides gas chromatograph/mass spectrometer (GC/MS)
criteria   appropriate   for  the   qualitative   confirmation   of   compound
identifications.

      1.4   Only   experienced   analysts   should   be   allowed  to  work   with
diazomethane due to the  potential  hazards associated with its use (the  compound
is explosive and carcinogenic).


2.0   SUMMARY OF METHOD

      2.1   Method   8150   provides   extraction,    esterification,   and   gas
chromatographic conditions  for  the  analysis  of   chlorinated  acid  herbicides.
Spiked samples  are used  to verify the applicability  of  the  chosen  extraction
technique to each  new  sample  type.   The esters are  hydrolyzed with  potassium


                                  8150B -  1                         Revision 2
                                                                September 1994

-------
hydroxide, and extraneous organic material  is removed by a  solvent wash.  After
acidification, the acids  are extracted with  solvent and converted to their methyl
esters using diazomethane  as  the  derivatizing agent.  After excess  reagent is
removed, the esters are determined by gas chromatography employing an electron
capture  detector,    microcoulometric detector,  or  electrolytic  conductivity
detector  (Goerlitz  and Lamar,  1967).   The  results  are reported as  the acid
equivalents.

      2.2   The  sensitivity  of Method  8150  usually  depends  on the  level  of
interferences rather than on instrumental limitations.
3.0   INTERFERENCES

      3.1   Refer to Method 8000.

      3.2   Method  interferences  may be  caused  by contaminants  in solvents,
reagents, glassware, and other sample processing  hardware that lead  to discrete
artifacts or elevated baselines in gas chromatograms.  All these materials must
be routinely demonstrated to be free  from  interferences under the conditions of
the analysis, by analyzing reagent blanks.

            3.2.1 Glassware must  be  scrupulously cleaned.  Clean each piece of
      glassware  as  soon as  possible after  use  by rinsing  it  with  the  last
      solvent used in it.  This should be followed by detergent washing with hot
      water and  rinses  with tap water,  then with  organic-free  reagent water.
      Glassware  should  be  solvent-rinsed  with  acetone  and  pesticide-quality
      hexane.  After rinsing and drying, glassware should be sealed and stored
      in  a  clean environment  to  prevent  any  accumulation  of  dust  or  other
      contaminants.    Store  glassware inverted  or capped with  aluminum  foil.
      Immediately prior to use, glassware should be rinsed with the next solvent
      to be used.

            3.2.2 The use of high  purity reagents and solvents helps  to minimize
      interference problems.  Purification of solvents by distillation in all-
      glass systems may be required.

      3.3   Matrix  interferences  may  be  caused   by  contaminants  that  are
coextracted from the sample.   The  extent  of  matrix  interferences  will  vary
considerably from waste  to waste, depending upon the nature and diversity of the
waste being sampled.

      3.4   Organic acids,  especially chlorinated acids,  cause  the most direct
interference with the determination.  Phenols, including chlorophenols, may also
interfere with this procedure.

      3.5   Alkaline hydrolysis and subsequent  extraction of the basic solution
remove many chlorinated hydrocarbons  and phthalate esters that  might otherwise
interfere with the electron  capture analysis.

      3.6   The  herbicides,  being strong  organic  acids,  react  readily  with
alkaline substances and may  be lost during analysis.  Therefore, glassware and
glass wool  must be  acid rinsed,  and sodium  sulfate must  be  acidified  with
sulfuric acid prior to use to avoid this possibility.

                                   8150B -  2                         Revision 2
                                                                September 1994

-------
      3.7   Sample  extracts  should be dry  prior to methylation  or  else poor
recoveries will be obtained.


4.0   APPARATUS AND MATERIALS

      4.1   Gas chromatograph

            4.1.1 Gas  chromatograph,  analytical  system  complete  with  gas
      chromatograph  suitable  for   on-column   injections   and  all   required
      accessories, including  detectors, analytical columns,  recorder, gases, and
      syringes.  A data system for measuring peak heights and/or peak areas is
      recommended.

            4.1.2 Columns

                  4.1.2.1     Column  la and Ib - 1.8 m x 4 mm ID glass, packed
            with  1.5%  SP-2250/1.95%  SP-2401  on Supelcoport  (100/120  mesh)  or
            equivalent.

                  4.1.2.2     Column  2 - 1.8 m x 4 mm ID glass, packed with 5%
            OV-210 on Gas Chrom Q (100/120 mesh) or equivalent.

                  4.1.2.3     Column  3  -  1.98  m x 2 mm  ID  glass,  packed with
            0.1% SP-1000 on 80/100 mesh Carbopack C or equivalent.

            4.1.3 Detector - Electron capture (ECD).

      4.2   Erlenmeyer flasks - 250  and 500 ml  Pyrex,  with  24/40 ground glass
joint.

      4.3   Beaker - 500 ml.

      4.4   Diazomethane generator -  Refer to Sec. 7.4 to determine which method
of diazomethane generation should be  used  for a particular application.

            4.4.1 Diazald kit - recommended for the generation of diazomethane
      using the procedure given in Sec. 7.4.2  (Aldrich Chemical  Co.,  Cat.  No.
      210,025-2 or equivalent).

            4.4.2 Assemble from two 20 x 150 mm test tubes, two Neoprene rubber
      stoppers, and a  source of  nitrogen.   Use Neoprene rubber  stoppers with
      holes drilled in  them to  accommodate glass delivery tubes.  The exit tube
      must  be  drawn to  a point  to   bubble  diazomethane  through the  sample
      extract.   The generator assembly is shown in Figure 1.  The procedure for
      use of this type  of generator is given  in Sec.  7.4.3.

      4.5   Vials - 10 to  15 ml,  amber glass,  with Teflon  lined  screw cap  or
crimp top.

      4.6   Separatory  funnel - 2000  ml,  125 ml,  and 60 ml.

      4.7   Drying column - 400 mm x  20 mm ID Pyrex chromatographic column with
Pyrex glass wool at bottom and a  Teflon stopcock.

                                  8150B - 3                         Revision 2
                                                                September 1994

-------
      NOTE:  Fritted  glass  discs are  difficult  to decontaminate  after highly
            contaminated extracts  have been passed through.   Columns without
            frits may  be  purchased.   Use  a small  pad of Pyrex  glass wool  to
            retain the  adsorbent.   Prewash  the  glass wool  pad with  50  mL  of
            acetone  followed by 50 mL  of elution solvent prior to packing the
            column with adsorbent.

      4.8   Kuderna-Danish (K-D) apparatus

            4.8.1 Concentrator tube -  10 ml, graduated (Kontes K-570050-1025 or
      equivalent).   A  ground  glass stopper  is used to prevent evaporation  of
      extracts

            4.8.2 Evaporation   flask   -   500   mL  (Kontes   K-570001-500  or
      equivalent).   Attach  to  concentrator  tube with springs,  clamps  or
      equivalent.

            4.8.3 Snyder column  -  Three  ball  macro  (Kontes  K-503000-0121  or
      equivalent).

            4.8.4 Snyder  column -  Two  ball  micro  (Kontes  K-569001-0219  or
      equivalent).

            4.8.5 Springs -  1/2 inch  (Kontes K-662750 or equivalent).

      4.9   Boiling chips - Solvent extracted, approximately  10/40 mesh (silicon
carbide or equivalent).

      4.10  Water  bath  -  Heated,  with  concentric  ring   cover,  capable  of
temperature control   (+ 5°C).   The bath  should be  used  in  a hood.

      4.11  Microsyringe - 10 /xL.

      4.12  Wrist shaker - Burrell  Model 75 or equivalent.

      4.13  Glass wool  - Pyrex, acid washed.

      4.14  Balance  - Analytical,  capable of accurately weighing to 0.0001 g.

      4.15  Syringe  - 5 mL.

      4.16  Glass rod.


5.0   REAGENTS

      5.1   Reagent  grade  inorganic  chemicals   shall   be used  in all  tests.
Unless otherwise indicated, it is intended that all reagents shall  conform to the
specifications of the Committee on Analytical Reagents  of the American Chemical
Society, where  such  specifications  are available.   Other grades  may  be  used,
provided it  is first  ascertained that the reagent  is of sufficiently high purity
to permit its use without  lessening the accuracy of the determination.
                                  8150B  - 4                         Revision 2
                                                                September 1994

-------
      5.2   Organic-free reagent water.  All references to water in this  method
refer to organic-free reagent water, as defined  in  Chapter One.

      5.3   Sulfuric acid solution

            5.3.1  ((1:1) (v/v)) - Slowly add 50 mL H2S04 (sp. gr. 1.84)  to 50  ml
      of organic-free reagent water.

            5.3.2  ((1:3) (v/v)) - Slowly add 25 ml H2S04 (sp. gr. 1.84)  to 75  ml
      of organic-free reagent water.

      5,4   Hydrochloric  acid   ((1:9)   (v/v)),   HC1.     Add  one   volume   of
concentrated HC1 to 9 volumes of organic-free  reagent water.

      5.5   Potassium hydroxide  solution (KOH)  -  37%  aqueous solution  (w/v).
Dissolve 37 g  potassium hydroxide pellets  in  organic-free reagent water, and
dilute to 100 mL.

      5.6   Carbitol (Diethylene glycol  monoethyl  ether), C2H5OCH2CH2OCH2CH2OH.
 Available from Aldrich Chemical Co.

      5.7   Solvents

            5.7.1  Acetone, CH3COCH3 - Pesticide quality or equivalent.

            5.7.2  Methanol, CH3OH - Pesticide  quality or  equivalent.

            5.7.3  Isooctane,   (CH3)3CCH2CH(CH3)2   -   Pesticide   quality   or
      equivalent.

            5.7.4  Hexane, C6H14 -  Pesticide  quality or  equivalent.

            5.7.5  Diethyl Ether,  C2H5OC2H5.    Pesticide  quality or  equivalent.
      Must  be  free of  peroxides as  indicated by  test   strips  (EM Quant,  or
      equivalent).  Procedures  for  removal  of peroxides  are provided  with the
      test strips.  After cleanup, 20 ml of ethyl alcohol  preservative must  be
      added to each liter of ether.

      5.8   Sodium sulfate (granular,  acidified,  anhydrous),  Na2S04. Purify  by
heating at  400°C for 4  hours  in a shallow  tray,  or by precleaning the  sodium
sulfate with  methylene  chloride.   If the  sodium  sulfate  is  precleaned with
methylene chloride, a method blank must be analyzed, demonstrating that  there  is
no interference  from  the sodium sulfate.  'Acidify by  slurrying  100  g  sodium
sulfate with enough diethyl ether  to just  cover the solid; then add 0.1  ml  of
concentrated sulfuric  acid and  mix thoroughly.  Remove the ether under a vacuum.
Mix 1  g  of the  resulting  solid with  5  ml  of organic-free  reagent water and
measure the pH of the mixture.  It must be below a  pH of  4.  Store  at  130°C.

      5.9   N-Methyl-N-nitroso-p-toluenesulfonamide (Diazald), CH3C6H4S02N(CH3)NO.
 Available from Aldrich Chemical Co.

      5.10  Silicic acid.  Chromatographic  grade,  nominal 100 mesh.   Store  at
130°C.

                                   8150B  -  5                         Revision 2
                                                                September 1994

-------
      5.11  Stock standard solutions -  Stock standard  solutions can be prepared
from pure standard materials or purchased as certified solutions.

            5.11.1      Prepare stock standard solutions by accurately weighing
      about 0.0100 g of pure  acids.   Dissolve the  acids  in  pesticide quality
      acetone and dissolve the  esters in  10% acetone/isooctane (v/v) and dilute
      to volume in a  10 mL volumetric flask.   Larger volumes can be used at the
      convenience of the  analyst.   If  compound purity is certified  at  96% or
      greater, the  weight  can  be  used  without  correction  to  calculate  the
      concentration  of  the stock  standard.    Commercially  prepared  stock
      standards can be used at any concentration  if they  are  certified  by the
      manufacturer or by an independent source.

            5.11.2      Transfer the stock  standard solutions  into vials  with
      Teflon  lined screw  caps  or crimp  tops.   Store  at  4°C and  protect  from
      light.  Stock standard solutions should be checked frequently  for signs of
      degradation or evaporation, especially just prior to preparing calibration
      standards from them.

            5.11.3      Stock standard solutions of the derivatized acids  must
      be replaced after 1 year, or  sooner,  if comparison  with  check standards
      indicates a problem.  Stock standard solutions of the  free acids degrade
      more  quickly  and should  be  replaced  after  two  months,  or  sooner  if
      comparison with check standards indicates a problem.

      5.12  Calibration standards - A minimum of five calibration  standards  for
each parameter of interest should be  prepared through  dilution of  the stock
standards  with  diethyl  ether.    One  of the  concentrations  should  be  at  a
concentration  near,  but  above,  the method detection limit.    The  remaining
concentrations should correspond to the expected range of concentrations found
in real  samples  or should  define  the  working range  of  the GC.   Calibration
solutions must be replaced after  six months,  or sooner if comparison with check
standards indicates a problem.

      5.13  Internal  standards (if internal  standard calibration is used)  - To
use this approach, the analyst  must select one or more internal  standards  that
are similar in analytical  behavior to  the compounds of  interest.   The analyst
must further  demonstrate  that the measurement of  the  internal  standard  is not
affected by method or matrix interferences.   Because  of  these  limitations,  no
internal standard can be suggested that is applicable  to all  samples.

            5.13.1      Prepare  calibration  standards  at a  minimum of  five
      concentrations  for each parameter of interest as described in Sec. 5.12.

            5.13.2      To  each  calibration  standard,  add  a  known  constant
      amount  of  one  or  more internal  standards,  and dilute  to  volume  with
      hexane.

            5.13.3      Analyze each calibration standard per Sec. 7.0.

      5.14  Surrogate standards - The analyst should monitor the performance of
the extraction, cleanup (when used),  and analytical system,  and the effectiveness
of the  method in dealing  with  each  sample matrix  by  spiking each  sample,


                                  8150B  - 6                        Revision 2
                                                                September  1994

-------
standard,  and  organic-free  reagent water  blank  with one  or  two  herbicide
surrogates (e.g.  herbicides that are not  expected  to  be present  in the sample).
The surrogates selected should elute over the range of the temperature program
used in this method.  2,4-Dichlorophenylacetic acid  (DCAA) is recommended as a
surrogate  compound.   Deuterated  analogs of  analytes  should  not be  used  as
surrogates for gas chromatographic analysis due to coelution problems.


6.0   SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1   See the  introductory  material  to this Chapter,  Organic  Analytes,
Sec. 4.1.  Extracts  must be  stored  under refrigeration and analyzed  within 40
days of extraction.


7.0  PROCEDURE

      7.1   Preparation of waste samples

            7.1.1  Extraction

                   7.1.1.1     Follow Method 3580 except use diethyl ether as the
            dilution solvent, acidified anhydrous  sodium sulfate,  and acidified
            glass wool.

                   7.1.1.2     Transfer 1.0 mL  (a lesser volume or a dilution may
            be required if  herbicide concentrations are high)  to  a 250 mL ground
            glass-stoppered Erlenmeyer  flask.  Proceed to Sec. 7.2.2 hydrolysis.

      7.2   Preparation of soil, sediment, and other solid samples

            7.2.1  Extraction

                   7.2.1.1     To a  500 mL,  wide mouth  Erlenmeyer flask add 50
            g (dry weight as  determined in Method  3540, Sec. 7.2.1) of the well
            mixed, moist solid  sample.   Adjust  the pH to  2  (See Method  9045)
            with  concentrated  HC1  and monitor  the   pH  for  15  minutes  with
            occasional stirring.  If necessary, add additional HC1 until the pH
            remains at 2.

                  7.2.1.2     Add  20  mL  acetone  to  the  flask  and  mix  the
            contents with the wrist shaker  for 20 minutes.  Add 80 mL diethyl
            ether to the same flask  and shake  again for 20 minutes.  Decant the
            extract and measure the volume of solvent recovered.

                  7.2.1.3     Extract  the  sample  twice  more using  20 mL  of
            acetone followed  by 80 mL of diethyl ether.  After addition of each
            solvent, the mixture  should  be shaken with  the  wrist shaker  for
            10 minutes and  the acetone-ether extract  decanted.

                  7.2.1.4     After the third  extraction, the volume of extract
            recovered should be at  least  75%  of  the volume of  added  solvent.
            If this is not the  case, additional extractions may be necessary.
            Combine  the  extracts  in a 2 liter  separatory funnel containing

                                  8150B - 7                        Revision  2
                                                                September  1994

-------
250 ml of reagent water.   If  an  emulsion  forms,  slowly add 5 g of
acidified sodium sulfate (anhydrous) until  the solvent-water mixture
separates.   A  quantity of  acidified  sodium sulfate equal  to the
weight of the sample may be added, if necessary.

      7.2.1.5     Check the pH of the extract.  If it is not at or
below pH  2, add more concentrated  HC1  until  stabilized  at the
desired pH.  Gently mix the contents  of the separatory funnel for
1 minute  and allow  the layers to  separate.  Collect  the aqueous
phase in  a clean beaker  and  the extract  phase  (top layer)  in  a
500 mL ground glass-stoppered Erlenmeyer flask.   Place the aqueous
phase back into the  separatory funnel  and  re-extract using 25 ml of
diethyl  ether.   Allow the layers to separate and discard the aqueous
layer.  Combine the ether extracts in the 500 ml Erlenmeyer flask.

      7.2.1.6     An   alternative   extraction   procedure   using
ultrasonic extraction can be found in Sec. 7.2 of Method 8151.

7.2.2 Hydrolysis

      7.2.2.1     Add 30 ml of organic-free reagent water, 5 ml of
37% KOH, and one or two clean  boiling chips to the flask.   Place a
three ball Snyder column on the flask, evaporate the diethyl ether
on a water bath, and continue to heat until the  hydrolysis step is
completed (usually 1 to 2 hours).

      7.2.2.2     Remove the flask from the water bath and allow to
cool.  Transfer the water solution to a 125 ml separatory funnel and
extract the  basic  solutions once with  40 ml and then  twice with
20 ml of  diethyl ether.   Allow sufficient time for  the  layers to
separate and discard the ether layer  each time.   The phenoxy-acid
herbicides remain soluble in the aqueous phase as  potassium salts.

7.2.3 Solvent cleanup

      7.2.3.1     Adjust  the  pH to  2  by  adding  5 ml  cold  (4°C)
sulfuric acid (1:3)  to  the  separatory  funnel.  Be sure to check the
pH at this point.  Extract  the herbicides  once with 40 mL and twice
with 20 ml of diethyl  ether.  Discard the aqueous phase.

      7.2.3.2     Combine  ether  extracts  in  a  125  ml  Erlenmeyer
flask  containing 5-7  g of  acidified  anhydrous  sodium  sulfate.
Stopper  and allow  the  extract  to  remain  in  contact  with  the
acidified sodium sulfate.   If concentration and  esterification are
not to be performed immediately,  store the sample  overnight in the
refrigerator.

      NOTE: The  drying step is very  critical  to  ensuring  complete
            esterification.   Any moisture remaining in  the ether
            will result in low herbicide recoveries. The amount of
            sodium  sulfate   is   adequate  if   some  free=flowing
            crystals are visible when  swirling  the  flask.   If all
            the  sodium sulfate  solidifies  in  a  cake,  add  a few
                      8150B  -  8                         Revision 2
                                                    September 1994

-------
                  additional grams of acidified sodium sulfate and again
                  test by swirling.  The 2 hour drying time  is a minimum,
                  however, the extracts may be  held overnight in contact
                  with the sodium  sulfate.

            7.2.3.3     Transfer  the ether  extract,  through a  funnel
      plugged  with  acid  washed  glass  wool,   into  a 500  ml K-D  flask
      equipped with a 10 ml concentrator tube.  Use a glass rod to crush
      caked sodium  sulfate during the  transfer.   Rinse  the Erlenmeyer
      flask and  column  with 20-30 ml of diethyl  ether to  complete the
      quantitative transfer.

            7.2.3.4     Add one or two  clean  boiling  chips  to the flask
      and attach a three ball  Snyder  column.  Prewet the Snyder column by
      adding about 1 ml  of diethyl  ether to the  top.  Place the apparatus
      on  a  hot water bath  (60°-65°C) so  that  the concentrator  tube is
      partially  immersed  in  the  hot  water and  the entire  lower  rounded
      surface  of the flask  is bathed  in vapor.    Adjust  the  vertical
      position of the apparatus and  the water  temperature,  as required,
      to complete the concentration in 15-20 minutes.  At the proper rate
      of distillation the balls of the column will actively chatter, but
      the chambers  will  not  flood.  When the  apparent  volume of liquid
      reaches 1 ml,  remove the K-D  apparatus from the water bath and allow
      it to drain and cool for at  least  10 minutes.

            7.2.3.5     Remove the Snyder column and rinse the flask and
      its lower joints into the concentrator tube  with 1-2 ml of diethyl
      ether.   A  5 mL  syringe is  recommended for this operation.   Add  a
      fresh boiling  chip,  attach a micro  Snyder  column to the concentrator
      tube, and prewet the column by  adding 0.5 ml of ethyl  ether to the
      top.  Place the micro K-D apparatus  on the  water  bath so  that the
      concentrator tube is partially immersed  in  the  hot  water.   Adjust
      the vertical  position of the  apparatus and the water temperature as
      required  to  complete  concentration in  5-10 minutes.   When  the
      apparent volume of the liquid reaches 0.5 ml, remove the micro K-D
      from the  bath  and  allow it  to  drain and  cool.  Remove the Snyder
      column  and add  0.1  ml  of  methanol.     Rinse  the  walls  of  the
      concentrator tube  while adjusting  the extract volume to 1.0 ml with
      diethyl  ether.  Proceed to Sec. 7.4 for esterification.

7.3   Preparation of aqueous samples

      7.3.1 Extraction

            7.3.1.1     Using  a  1  liter  graduated cylinder, measure  1
      liter (nominal) of sample,  record  the sample volume  to the  nearest
      5 ml, and  quantitatively transfer it to  the  separatory funnel. If
      high concentrations are anticipated, a smaller  volume  may  be used
      and then diluted with organic-free reagent water  to 1 liter.  Adjust
      the pH to less than 2. with sulfuric acid (1:1).

            7.3.1.2     Add 150 ml  of diethyl ether to the sample  bottle,
      seal, and  shake for  30  seconds to rinse the walls.   Transfer the
                            8150B  - 9                         Revision 2
                                                          September 1994

-------
solvent wash  to  the separatory  funnel  and extract the  sample by
shaking the funnel  for 2 minutes  with  periodic  venting to release
excess pressure.   Allow the organic layer to separate from the water
layer  for  a  minimum  of  10  minutes.    If  the  emulsion  interface
between layers is  more than one third the size  of the solvent layer,
the analyst must  employ mechanical  techniques  to complete the phase
separation.  The  optimum technique depends  upon  the sample and may
include stirring,  filtration  of the emulsion through  glass  wool,
centrifugation, or other  physical  methods.  Drain the aqueous phase
into a 1 liter Erlenmeyer flask.  Collect the solvent extract in a
250 ml ground  glass Erlenmeyer flask containing 2 ml  of  37% KOH.
Approximately 80 ml of the diethyl  ether will remain  dissolved in
the aqueous phase.

      7.3.1.3     Repeat the extraction two more times using 50 ml
of diethyl  ether  each  time.  Combine the extracts in the Erlenmeyer
flask.  (Rinse the  1  liter flask with  each additional  aliquot of
extracting solvent.)

7.3.2 Hydrolysis

      7.3.2.1     Add one or  two  clean  boiling  chips  and  15  ml of
organic-free reagent water to  the  250 ml flask  and attach a  three
ball Snyder column.  Prewet the Snyder column  by adding about 1 ml
of diethyl  ether  to the  top of the column.   Place the  apparatus on
a hot water bath (60°-65°C) so that the bottom of  the flask is bathed
with hot water vapor.  Although the diethyl  ether will  evaporate in
about  15 minutes,  continue heating  until  the  hydrolysis  step is
completed   (usually  1  to  2  hours).    Remove the  apparatus  and let
stand at room temperature for at least 10 minutes.

      7.3.2.2     Transfer the solution  to a 60 ml separatory funnel
using  5-10 ml of  organic-free  reagent water.    Wash the  basic
solution twice  by  shaking  for 1  minute with  20  ml  portions  of
diethyl ether.  Discard the organic  phase.  The herbicides remain
in the aqueous phase.

7.3.3 Solvent cleanup

      7.3.3.1     Acidify the  contents of the separatory funnel to
pH 2 by adding 2 ml  of cold (4°C) sulfuric acid (1:3).  Test with pH
indicator  paper.   Add 20  ml diethyl ether and  shake vigorously for
2 minutes.   Drain  the  aqueous,layer  into  a 250 ml Erlenmeyer flask,
and pour the organic layer into a 125 mL  Erlenmeyer flask containing
about  5-7  g of acidified sodium  sulfate.   Repeat the extraction
twice  more  with   10 mL  aliquots of  diethyl  ether, combining all
solvent in  the 125 ml flask.  Allow the extract to remain in contact
with the sodium sulfate  for approximately 2 hours.

      NOTE: The drying step is  very  critical  to ensuring  complete
            esterification.   Any  moisture  remaining  in the  ether
            will  result  in  low herbicide recoveries.  The amount of
            sodium  sulfate  is  adequate  if  some  free   flowing


                      8150B -  10                        Revision 2
                                                    September 1994

-------
                  crystals  are  visible  when  swirling the flask.  If all
                  the  sodium sulfate  solidifies in  a  cake, add  a few
                  additional grams of acidified  sodium sulfate  and again
                  test by swirling.  The 2 hour drying time  is a minimum,
                  however,  the  extracts may be held  overnight in contact
                  with the  sodium  sulfate.

            7.3.3.2     Transfer  the  ether  extract,  through  a  funnel
      plugged  with  acid  washed glass  wool,   into  a 500  ml K-D  flask
      equipped with a 10 mL concentrator tube.   Use  a glass rod to crush
      caked sodium  sulfate during the  transfer.   Rinse  the Erlenmeyer
      flask and  column  with 20-30 mL  of diethyl ether  to  complete the
      quantitative transfer.

            7.3.3.3     Add one or two  clean  boiling chips to the flask
      and attach a three ball Snyder  column.  Prewet  the Snyder column by
      adding about 1 mL  of diethyl  ether to the top.   Place the apparatus
      on  a  hot water bath  (60°-65°C) so  that  the  concentrator tube is
      partially  immersed  in the hot  water and  the  entire  lower rounded
      surface  of the flask is bathed  in vapor.    Adjust  the  vertical
      position of the apparatus and  the water  temperature,  as required,
      to complete the concentration in  15-20 minutes.  At the proper rate
      of distillation the balls of the column will actively chatter, but
      the chambers  will  not flood.  When  the  apparent  volume of liquid
      reaches 1 mL,  remove the K-D  apparatus from the water bath  and allow
      it to drain and cool  for  at  least 10 minutes.

            7.3.3.4     Remove  the Snyder  column and  rinse the flask and
      its lower joints into the concentrator tube with 1-2 mL of diethyl
      ether.   A  5 mL  syringe is recommended for this operation.   Add a
      fresh boiling  chip,  attach a micro Snyder column to the concentrator
      tube, and prewet the  column by adding 0.5 mL of ethyl  ether to the
      top.  Place the micro K-D apparatus  on  the water  bath so  that the
      concentrator tube is  partially immersed  in the hot water.  Adjust
      the vertical  position of the  apparatus and  the  water temperature as
      required  to  complete  concentration in  5-10  minutes.    When  the
      apparent volume of the liquid reaches 0.5 mL, remove the micro K-D
      from the bath  and  allow it to drain and  cool.   Remove the Snyder
      column  and add  0.1  mL  of  methanol.    Rinse the  walls of  the
      concentrator tube while adjusting  the extract  volume to 1.0 mL with
      diethyl  ether.

7.4   Esterification

      7.4.1 Two methods may be  used  for the  generation  of diazomethane:
the bubbler method  (set  up shown in Figure  1) and the  Diazald kit method.
The bubbler method  is suggested when  small  batches  (10-15)  of samples
require esterification.   The bubbler method works well with samples that
have low concentrations  of herbicides (e.g. aqueous  samples) and is safer
to use than  the Diazald kit  procedure.  The Diazald kit method  is good for
large quantities  of samples  needing esterification.   The  Diazald  kit
method is more effective than the bubbler method for soils or samples that
may contain  high concentrations  of herbicides  (e.g.,  samples  such as soils


                            8150B - 11                         Revision 2
                                                          September 3994

-------
that result in yellow extracts  following  hydrolysis  may be difficult to
handle by the bubbler method).  The diazomethane derivatization  (U.S. EPA,
1971) procedures,  described below,  will  react efficiently with  all of the
chlorinated herbicides described in this method and should be used only by
experienced analysts, due  to the potential hazards  associated with its
use.  The following precautions should be taken:

      CAUTION:     Diazomethane  is a  carcinogen   and  can   explode  under
                  certain  conditions.

                  Use a safety  screen.
                  Use mechanical pipetting aides.
                  Do not heat above 90°C -- EXPLOSION may  result.
                  Avoid grinding  surfaces,  ground glass joints,  sleeve
                  bearings, glass stirrers --     EXPLOSION may result.
                  Store away from alkali metals -- EXPLOSION may result.
                  Solutions  of  diazomethane  decompose rapidly in  the
                  presence  of  solid  materials  such  as copper  powder,
                  calcium chloride, and boiling chips.

      7.4.2 Diazald kit method  - Instructions for preparing diazomethane
are provided with the generator kit.

            7.4.2.1     Add 2 mL of diazomethane  solution  and let sample
      stand for 10 minutes with occasional swirling.

            7.4.2.2     Rinse  inside  wall  of the  ampule  with  several
      hundred  /uL  of   diethyl   ether.     Allow   solvent   to  evaporate
      spontaneously at room temperature to about  2 mL.

            7.4.2.3     Dissolve the residue in 5 mL of hexane.  Analyze
      by gas chromatography.

      7.4.3 Bubbler  method  -   Assemble  the  diazomethane  bubbler  (see
Figure 1).

            7.4.3.1     Add 5 mL of diethyl  ether  to the first test tube.
      Add 1 mL of diethyl  ether, 1  mL  of carbitol, 1.5 mL of 37% KOH, and
      0.1-0.2 g Diazald to the  second  test  tube.   Immediately place the
      exit tube into  the concentrator tube containing  the sample extract.

            Apply nitrogen flow  (10 mL/min) to bubble diazomethane through
      the extract for 10 minutes or until the yellow color of diazomethane
      persists.    The   amount   of  Diazald  used   is   sufficient  for
      esterification  of  approximately   three   sample  extracts.     An
      additional 0.1-0.2  g  of  Diazald  may  be added  (after  the initial
      Diazald is consumed) to extend the generation of the  diazomethane.
      There is  sufficient  KOH present  in the original solution to perform
      a maximum of approximately 20 minutes of total  esterification.

            7.4.3.2     Remove  the concentrator  tube  and seal  it  with a
      Neoprene or Teflon stopper.  Store at room temperature in  a hood for
      20 minutes.
                            8150B - 12                        Revision 2
                                                          September 1994

-------
                  7.4.3.3     Destroy  any  unreacted  diazomethane  by  adding
            0.1-0.2 g  silicic acid  to  the concentrator  tube.   Allow to stand
            until the evolution  of nitrogen gas has stopped.  Adjust  the sample
            volume to  10.0 mi with  hexane.   Stopper the concentrator tube and
            store refrigerated  if  further processing  will  not  be  performed
            immediately.   It is  recommended  that the  methylated extracts be
            analyzed immediately to minimize  the trans-esterification and other
            potential reactions  that may  occur.  Analyze  by gas chromatography.

      7.5   Gas chromatographic conditions (Recommended)

            7.5.1 Column la

            Carrier gas (5% methane/95% argon)  flow rate:    70 mL/min
            Temperature program:     185°C, isothermal.

            7.5.2 Column Ib

            Carrier gas (5% methane/95% argon)  flow rate:    70 mL/min
            Initial  temperature:     140°C, hold for 6 minutes
            Temperature program:     140°C to 200°C at  10°C/min, hold  until last
                                     compound  has  eluted.

            7.5.3 Column 2

            Carrier gas (5% methane/95% argon)  flow rate:    70 mL/min
            Temperature program:     185°C, isothermal.

            7.5.4 Column 3

            Carrier gas (ultra-high purity N2) flow rate:    25 mL/min
            Initial  temperature:     100°C, no  hold
            Temperature program:     100°C to 150°C at  10°C/min, hold  until last
                                     compound  has  eluted.

      7.6   Calibration  -   Refer   to  Method  8000  for  proper  calibration
techniques.  Use Table  1 and especially Table 2 for guidance on selecting the
lowest point on the  calibration  curve.

            7.6.1 The  procedure for internal  or external calibration  may be
      used.  Refer to Method 8.000 for a  description of  each of these procedures.

            7.6.2 The following  gas chromatographic columns are recommended for
      the compounds  indicated:


            Analyte           Column             Analyte            Column

            Dicamba             la,2              Dalapon              3
            2,4-D               la,2              MCPP                 Ib
            2,4,5-TP            la,2              MCPA                 Ib
            2,4,5-T             la,2              Dichloroprop         Ib
            2,4-DB              la                Dinoseb              Ib


                                  8150B -  13                        Revision 2
                                                                September 1994

-------
      7.7   Gas chromatographic analysis

            7.7.1 Refer to Method 8000.   If the  internal  standard calibration
      technique is used,  add 10 jiL of internal standard to the sample prior to
      injection.

            7.7.2 Method 8000  provides  instructions  on  the analysis sequence,
      appropriate dilutions,  establishing  daily  retention time  windows,  and
      identification criteria.  Include a mid-concentration check standard after
      each group of 10 samples in the analysis sequence.

            7.7.3 Examples  of chromatograms  for  various chlorophenoxy  acid
      herbicides are shown in Figures 2 through 4.

            7.7.4 Record the sample volume injected and the resulting peak sizes
      (in area units or peak heights).

            7.7.5 Using either the internal  or external calibration procedure
      (Method 8000), determine the identity  and quantity of each component peak
      in the  sample  chromatogram  which corresponds to  the compounds  used  for
      calibration purposes.

            7.7.6 If calibration  standards have been analyzed  in the same manner
      as the samples (e.g. have undergone hydrolysis and esterification),  then
      the calculation  of  concentration given  in  Method 8000 should  be used.
      However,  if calibration  is  done  using standards  made  from  methyl ester
      compounds (compounds not esterified by application of this  method),  then
      the  calculation  of  concentration  must  include  a  correction  for  the
      molecular weight of the methyl  ester versus the acid herbicide.

            7.7.7 If peak  detection  and  identification are  prevented  due  to
      interferences, further  cleanup is required.   Before using  any  cleanup
      procedure, the analyst  must process  a  series  of standards  through  the
      procedure to validate elution  patterns  and  the  absence  of  interferences
      from reagents.


8.0   QUALITY CONTROL

      8.1   Refer  to  Chapter  One for  specific  quality   control  procedures.
Quality control to validate sample extraction is  covered in Method 3500 and in
the extraction method utilized.  If extract cleanup was performed,  follow the QC
in Method 3600 and in the specific cleanup method.

      8.2   Procedures to  check  the  GC  system  operation  are found  in  Method
8000.

            8.2.1 Select a representative spike concentration for each compound
      (acid or ester) to  be measured.  Using stock standards,  prepare a quality
      control check sample concentrate  in  acetone  1,000 times more concentrated
      than the selected concentrations.
                                  8150B - 14                        Revision 2
                                                                September 1994

-------
            8.2.2 Table 3 indicates single operator accuracy and precision for
      this method.   Compare  the  results obtained  with  the results  given in
      Table 3 to determine if the data quality is acceptable.

      8.3   Calculate surrogate standard  recovery  on  all  samples,  blanks, and
spikes.   Determine  if  the  recovery is  within limits (limits  established by
performing QC procedures outlined in Method 8000).

            8.3.1 If recovery is not within limits,  the following procedures are
      required.

                  •     Check to  be sure  there are no errors in calculations,
                        surrogate  solutions  and  internal  standards.    Also,
                        check instrument  performance.

                  •     Recalculate the  data and/or reanalyze the  extract if
                        none of the above checks reveal a problem.

                  •     Re-extract  and  re-analyze the sample  if none  of the
                        above are a problem  or flag  the data  as  "estimated
                        concentration".

      8.4   GC/MS confirmation

            8.4.1 GC/MS techniques  should be judiciously employed  to  support
      qualitative identifications made with this method.   Refer to Method 8270
      for the appropriate GC/MS  operating conditions and  analysis procedures.

            8.4.2 When  available,  chemical   ionization  mass  spectra  may  be
      employed to aid the qualitative  identification process.

            8.4.3 Should  these  MS  procedures  fail   to  provide  satisfactory
      results, additional  steps may be taken before  reanalysis.  These steps may
      include the use of alternate packed  or capillary GC columns or additional
      cleanup.


9.0   METHOD PERFORMANCE

      9.1   In  a single  laboratory,   using  organic-free reagent  water  and
effluents from publicly owned treatment  works (POTW), the  average  recoveries
presented in  Table  3  were obtained.   The standard deviations of the  percent
recoveries of these  measurements  are also included  in  Table  3.
10.0  REFERENCES

1.     U.S.  EPA, National  Pollutant  Discharge Elimination System,  Appendix  A,
      Fed.  Reg.,  38,  No.  75,  Pt.  II,  Method  for  Chlorinated Phenoxy  Acid
      Herbicides in Industrial Effluents, Cincinnati,  Ohio,  1971.

2.     Goerlitz, D.G., and  W.L.  Lamar, "Determination of Phenoxy Acid Herbicides
      in Water  by  Electron  Capture and  Microcoulometric  Gas  Chromatography,"
      U.S.  Geol. Survey Water Supply Paper,  1817-C,  1967.

                                  8150B - 15                        Revision 2
                                                                September 1994

-------
Burke, J.A.,  "Gas  Chromatography for Pesticide  Residue  Analysis;   Some
Practical Aspects,"   Journal  of the Association  of Official  Analytical
Chemists, 48, 1037, 1965.

U.S.  EPA,  "Extraction  and  Cleanup Procedure  for the  Determination  of
Phenoxy Acid Herbicides  in  Sediment,"  EPA Toxicant and Analysis Center,
Bay St. Louis, Mississippi, 1972.

"Pesticide Methods  Evaluation,"  Letter Report #33 for EPA Contract No. 68-
03-2697.    Available   from   U.S.   Environmental   Protection   Agency,
Environmental Monitoring and Support Laboratory, Cincinnati, Ohio 45268.

Eichelberger, J.W., L.E. Harris, and W.L.  Budde,  "Reference Compound to
Calibrate   Ion   Abundance   Measurement   in    Gas   Chromatography-Mass
Spectrometry," Analytical Chemistry, 47, 995,  1975.

Glaser,  J.A.  et.al.,  "Trace  Analysis  for Wastewaters,"  Environmental
Science & Technology, 15, 1426,  1981.

Gurka, D.F, Shore,  F.L., Pan, S-T,  "Single Laboratory Validation of EPA
Method 8150  for Determination  of  Chlorinated  Herbicides  in  Hazardous
Waste", JAOAC, 69,  970,  1986.

U.S.  EPA,  "Method  615.  The Determination of Chlorinated  Herbicides  in
Industrial and Municipal  Wastewater," Environmental Monitoring and Support
Laboratory, Cincinnati, Ohio,  45268, June  1982.
                            8150B - 16                        Revision 2
                                                          September 1994

-------
                                   TABLE 1.
                CHROMATOGRAPHIC  CONDITIONS  AND DETECTION LIMITS
                           FOR CHLORINATED  HERBICIDES
Compound
     Retention time (min)a

Col.la   Col.lb  Col.2   Col.3
"Column  conditions  are given  in  Sees.  4.1  and  7.5.
                                   TABLE 2.
                    DETERMINATION  OF  ESTIMATED  QUANTITATION
                      LIMITS (EQL) FOR VARIOUS MATRICES8
  Method
 detection
limit (M9/L)
2,4-D
2,4-DB
2,4,5-T
2,4,5-TP (Silvex)
Dalapon
Dicamba
Dichloroprop
Dinoseb
MCPA
MCPP
2.0
4.1
3.4
2.7
-
1.2
-
-
-
-

-
-
-
-
-
4.8
11.2
4.1
3.4
1.6
-
2.4
2.0
5.0
1.0
-
-
-
-
1.2
0.91
0.20
0.17
5.8
0.27
0.65
0.07
249
192
    Matrix
                                Factor
Ground water (based on one liter sample si?e)
Soil/sediment and other solids
Waste samples
                                   10
                                  200
                              100,000
aEQL  = [Method detection limit (see Table  1)] X  [Factor  found  in  this table].
For non-aqueous samples, the factor is on a wet  weight basis.   Sample EQLs are
highly matrix  dependent.   The EQLs to  be determined herein are  provided for
guidance and may not always be achievable.
                                  8150B - 17
                                      Revision 2
                                  September 1994

-------
                                   TABLE 3.
                    SINGLE OPERATOR ACCURACY AND PRECISION'
Compound
2,4-D


Dalapon


2,4-DB


Dicamba


Dichlorprop


Dinoseb

MCPA


MCPP


2,4,5-T


2,4,5-TP


Sample
Type
DW
MW
MW
DW
MW
MW
DW
MW
MW
DW
MW
MW
DW
MW
MW
MW
MW
DW
MW
MW
DW
MW
MW
DW
MW
MW
DW
MW
MW
Spike
10.9
10.1
200
23.4
23.4
468
10.3
10.4
208
1.2
1.1
22.2
10.7
10.7
213
0.5
102
2020
2020
21400
2080
2100
20440
1.1
1.3
25.5
1.0
1.3
25.0
Mean
Recovery
75
77
65
66
96
81
93
93
77
79
86
82
97
72
100
86
81
98
73
97
94
97
95
85
83
78
88
88
72
Standard
deviation
4
4
5
8
13
9
3
3
6
7
9
6
2
3
2
4
3
4
3
2
4
3
2
6
4
5
5
4
5
aAll  results based upon seven replicate analyses.  Esterification performed using
the bubbler method.  Data obtained from reference 8.

DW = ASTM Type II
MW = Municipal water
                                  8150B - 18
    Revision 2
September 1994

-------
                             FIGURE  1.
                       DIAZOMETHANE GENERATOR
   nitrogen
             TIE
rubber stopper
                    U
                 lube 1
                                       a*
tube 2
                                                       glass tubing
                             8150B - 19
                     Revision 2
                 September 1994

-------
                       FIGURE  2.
      GAS  CHROMATOGRAM  OF CHLORINATED  HERBICIDES
Column: 1.5% SP-2250/1.95* SF-2401 on Suptteopoa (100/120 Mt*)
Ttmparatura: Isothermal at 18S°C
Dtttctor: Electron Capturt
          0     12346
            RETENTION TIME (MINUTES)
                       8150B - 20
    Revision  2
September  1994

-------
                  FIGURE  3.
GAS CHROMATOGRAM OF  CHLORINATED HERBICIDES
  Column: 1.5% SP-2260/1.95% SP-2401 on Suotlcoport (100/120
  Program: UO°C for 6 Min. 10°C/Minuti to 200°C
  Dtuctor: Electron Capturt
        468
        RETENTION TIME (MINUTES)
10
12
                 81508 - 21
             Revision  2
         September  1994

-------
                  FIGURE  4.
    GAS CHROMATOGRAM OF DALAPON, COLUMN 3
   J
          Column: 0.1% 9-1000 on 80/100 Mmh C*rtoop«k C
          Program: 100°C, 10°C/Min to 1BO°C
          Otnetor: Etoctron Capture
   0246
RETENTION TIME (MINUTES)
                  8150B - 22
    Revision  2
September  1994

-------
                                  METHOD  81 SOB
            CHLORINATED  HERBICIDES  BY GAS  CHROMATOGRAPHY
   7.2.1.1
Adjust lampla
 pH with HCI.
                                         Solid
                                        Sample
7.2.1.2 Extract
 sample with
 acetona and
 diethyl ethar.
7.2.1.3 Extract
  twica more.
    7.2.1.4
   Combine
   extract*.
 7.2.1.5 Check
 pH of extract,
    adjust if
   necessary.
Separata layer*.
  7.2.1.5
 Re-extract
and discard
  aqueous
  phase.
   7.1.1 Follow
 Method 3580 for
 extraction, u»mg
   diethyl ether,
acidified anhydrous
aodium sulfata and
  acidified glass
      wool.
7.2.2 Proceed
    with
 hydrolysis.
    7.1.1.2 Use
     1.0 mL of
    cample for
    hydrolysis.
7.2.3  Proceed
 with  solvent
   cleanup.
                                                              V
                                                   7.3.1.1 Adjust
                                                     sample pH
                                                    with H2S04.
 7.3.1.2 Extract
   with diethyl
      ether.
     7.3.1.3
Extract twica more,
   and combine
     extracts.
                                 7.3.2 Proceed
                                     with
                                  hydrolysis.
                                                    7.3.3 Proceed
                                                     with solvent
                                                       cleanup.
                                    8150B  -  23
                                                         Revision  2
                                                    September  1994

-------
                                       METHOD 81BOB
                                        (Continued)
7.4.3 Assemble
 diazomethane
   bubbler;
   generate
 diazomethane.
    7.4
  Choose
 method for
esterification
7.4.2 Prepare
diazomethane
 according to
     kit
 instructions.
                              7.5 Set
                          chromatographic
                            conditions.
                            7.6 Claibrate
                            according to
                           Method 8000.
                               I
                            7.6.2 Choose
                             appropriate
                             GC column.
                              7.7 Analyze
                              by GC (refer
                              to Method
                                8000).
                                7.7.7 Do
                              interferences
                              prevent peak
                               detection?
                              7.7.7 Process
                                series of
                                stendards
                              through system
                                cleanup.
                                         8150B -  24
                                                          Revision  2
                                                    September 1994

-------
00

-------
                                 METHOD  8151

   CHLORINATED HERBICIDES BY GC USING METHYLATION OR PENTAFLUOROBENZYLATION
                  DERIVATIZATION: CAPILLARY COLUMN TECHNIQUE


1.0   SCOPE AND APPLICATION

      1.1   Method 8151  is a  capillary  gas  chromatographic  (GC)  method  for
determining certain chlorinated acid herbicides and related compounds  in aqueous,
soil  and waste matrices.  Specifically, Method 8151 may  be  used to determine the
following compounds:
      Compound Name                                   CAS No."


      2,4-D                                           94-75-7
      2,4-DB                                          94-82-6
      2,4,5-TP (Silvex)                               93-72-1
      2,4,5-T                                         93-76-5
      Dalapon                                         75-99-0
      Dicamba                                       1918-00-9
      Dichloroprop                                   120-36-5
      Dinoseb                                         88-85-7
      MCPA                                            94-74-6
      MCPP                                            93-65-2
      4-Nitrophenol                                  100-02-1
      Pentachlorophenol                               87-86-5


      a      Chemical Abstract Services Registry Number.

      Because these compounds are produced and used in various forms (i.e., acid,
salt, ester,  etc.),  Method 8151  describes a hydrolysis step  that can be used to
convert  herbicide  esters into the acid form prior to analysis.  Herbicide esters
generally have a half-life of less than  one week in soil.

      1.2   When Method 8151  is  used to  analyze  unfamiliar samples,  compound
identifications  should be supported  by  at  least one  additional  qualitative
technique.   Sec.  8.4  provides  gas  chromatograph/mass spectrometer  (GC/MS)
criteria   appropriate   for    the    qualitative   confirmation   of   compound
identifications.

      1.3   The estimated detection limits for  each of the compounds in aqueous
and soil matrices  are  listed in  Table 1.   The  detection limits  for  a specific
waste sample may differ from those  listed,  depending  upon  the nature  of  the
interferences and the  sample matrix.
                                   8151 - 1                         Revision 0
                                                                September 1994

-------
      1.4   The following compounds may also be determined using this method:
      Compound Name                               CAS No."
      Acifluorfen                               50594-66-6
      Bentazon                                  25057-89-0
      Chloramben                                  133-90-4
      DCPA diacidb                               2136-79-0
      3,5-Dichlorobenzoic acid                     51-36-5
      5-Hydroxydicamba                           7600-50-2
      Picloram                                   1918-02-1
            Chemical Abstract Services Registry Number.

            DCPA monoacid and diacid metabolites  included in method scope; DCPA
            diacid metabolite used for validation studies.   DCPA is a dimethyl
            ester.
      1.5   This method  is  restricted to use  by or under  the  supervision of
analysts  experienced  in  the  use  of gas  chromatography and  skilled  in  the
interpretation of gas  chromatograms.   Each analyst must demonstrate the ability
to generate acceptable results with this method.

      1.6   Only experienced analysts should be  allowed to work with diazomethane
due to the potential hazards associated with its  use (explosive, carcinogenic).


2.0   SUMMARY OF METHOD

      2.1   Method  8151   provides   extraction,   derivatization,    and   gas
chromatographic conditions  for the analysis of chlorinated  acid herbicides in
water, soil, and waste samples.   An option for  the hydrolysis of esters is also
described.

            2.1.1 Water  samples  are  extracted  with  diethyl  ether  and  then
      esterified with  either  diazomethane  or  pentafluorobenzyl  bromide.   The
      derivatives are  determined  by gas chromatography with  an electron capture
      detector  (GC/ECD).  The results are reported as acid equivalents.

            2.1.2 Soil  and  waste  samples  are  extracted and  esterified  with
      either diazomethane or  pentafluorobenzyl bromide.  The  derivatives  are
      determined  by  gas chromatography  with  an  electron capture  detector
      (GC/ECD).  The results are reported as acid equivalents.

            2.1.3 If herbicide esters are to be determined  using this method,
      hydrolysis  conditions for the esters in water  and  soil  extracts  are
      described.
                                   8151 - 2                         Revision 0
                                                                September 1994

-------
      2.2   The sensitivity of Method 8151 depends on the level of interferences
in addition to  instrumental  limitations.   Table 1  lists  the  GC/ECD and GC/MS
detection limits that can be obtained in aqueous and soil matrices in  the absence
of interferences.  Detection  limits  for a typical  waste sample  should be higher.


3.0   INTERFERENCES

      3.1   Refer to Method 8000.

      3.2   Method  interferences  may be  caused  by contaminants  in solvents,
reagents, glassware, and other sample processing  hardware  that lead  to discrete
artifacts or elevated baselines in gas  chromatograms.  All these materials must
be routinely demonstrated to be free from  interferences  under the conditions of
the analysis,  by analyzing reagent blanks.

            3.2.1 Glassware must be scrupulously cleaned.  Clean each piece of
      glassware  as  soon as  possible after  use  by rinsing  it  with  the last
      solvent used in  it.  This should be followed by detergent washing with hot
      water and  rinses  with  tap water, then with  organic-free  reagent water.
      Glassware  should  be solvent-rinsed  with   acetone  and  pesticide-quality
      hexane.   After rinsing and drying, glassware  should be sealed and stored
      in  a  clean environment  to prevent  any  accumulation of  dust  or  other
      contaminants.   Store  glassware  inverted or  capped with  aluminum  foil.
      Immediately prior to use, glassware should be  rinsed with the next solvent
      to be used.

            3.2.2 The use of high purity reagents  and solvents helps  to minimize
      interference problems.  Purification of solvents by distillation in all-
      glass systems may be required.

      3.3   Matrix  interferences  may   be  caused   by   contaminants  that  are
coextracted from  the  sample.   The extent of  matrix interferences will  vary
considerably from waste  to waste, depending upon the nature and diversity of the
waste being sampled.

      3.4   Organic acids,  especially chlorinated acids, cause the most direct
interference  with  the   determination   by  methylation.    Phenols,  including
chlorophenols, may also  interfere with  this procedure.  The determination using
pentafluorobenzylation is more sensitive,  and more  prone to  interferences from
the presence of organic acids or phenols than by  methylation.

      3.5   Alkaline hydrolysis and  subsequent extraction  of the basic solution
removes many chlorinated hydrocarbons and phthalate esters that might otherwise
interfere with the electron capture analysis.  However, hydrolysis may result in
the loss  of dinoseb and the  formation  of  aldol  condensation products  if  any
residual acetone remains from the extraction  of solids.

      3.6   The  herbicides,  being  strong  organic   acids,  react readily  with
alkaline substances and  may  be lost  during analysis.  Therefore,  glassware must
be acid-rinsed and then rinsed to constant pH with  organic-free  reagent water.
Sodium sulfate must be acidified.
                                   8151 - 3                         Revision 0
                                                                September 1994

-------
      3.7   Sample extracts  should be dry  prior to methylation  or else poor
recoveries will be obtained.


4.0   APPARATUS AND MATERIALS

      4.1   Gas chromatograph

            4.1.1 Gas  chromatograph  -  Analytical  system  complete with  gas
      chromatograph suitable for  Grob-type  injection  using capillary columns,
      and  all  required accessories  including detector,  capillary analytical
      columns, recorder, gases, and syringes. A data system for measuring peak
      heights and/or peak areas is recommended.

            4.1.2 Columns

                  4.1.2.1     Narrow Bore Columns

                        4.1.2.1.1   Primary  Column  1  -  30  m x  0.25  mm,  5%
                  phenyl/95%  methyl   silicone  (DB-5,  J&W   Scientific,   or
                  equivalent),  0.25 p.m film thickness.

                        4.1.2.1.2   Primary Column la  (GC/MS)  - 30  m x 0.32 mm,
                  5%  phenyl/95%  methyl  silicone,   (DB-5,  J&W Scientific,  or
                  equivalent),  1  ^m film thickness.

                        4.1.2.1.3   Column  2  -  30  m  x  0.25  mm  DB-608  (J&W
                  Scientific or equivalent)  with a 25 /Ltm film thickness.

                        4.1.2.1.4   Confirmation Column -  30 m x  0.25  mm,  14%
                  cyanopropyl phenyl  silicone,  (DB-1701,  J&W Scientific,  or
                  equivalent),  0.25 /xm film thickness.

                  4.1.2.2     Wide-bore Columns

                        4.1.2.2.1   Primary Column - 30 m x 0.53 mm DB-608 (J&W
                  Scientific or equivalent)  with 0.83  /urn film thickness.

                        4.1.2.2.2   Confirmation Column -  30 m x  0.53  mm,  14%
                  cyanopropyl phenyl  silicone,  (DB-1701,  J&W Scientific,  or
                  equivalent),  1.0 jum film thickness.

            4.1.3 Detector - Electron Capture Detector (ECD)

      4.2   Kuderna-Danish (K-D)  apparatus

            4.2.1 Concentrator tube - 10 mL  graduated  (Kontes K-570050-1025 or
      equivalent).   A  ground glass  stopper  is used to prevent evaporation  of
      extracts.

            4.2.2 Evaporation  flask   -   500   mL   (Kontes  K-570001-500   or
      equivalent).    Attach  to  concentrator  tube  with  springs,  clamps,   or
      equivalent.


                                   8151 -  4                          Revision 0
                                                               September 1994

-------
            4.2.3 Snyder column  -  Three  ball  macro  (Kontes  K-503000-0121  or
      equivalent).

            4.2.4 Snyder column  -  Two  ball  micro  (Kontes  K-569001-0219  or
      equivalent).

            4.2.5 Springs - 1/2 inch (Kontes K-662750 or equivalent).

      4.3   Diazomethane Generator:  Refer to Sec. 7.5 to determine which method
of diazomethane generation  should be used for a particular generation.

            4.3.1 Diazald Kit - Recommended for the generation of diazomethane
      (Aldrich Chemical Co.,  Cat  No. 210,025-0, or equivalent).

            4.3.2 As an  alternative,  assemble  from  two  20 mm  x 150  mm  test
      tubes,  two  Neoprene  rubber  stoppers,  and  a source  of nitrogen.   Use
      Neoprene rubber stoppers with holes drilled in them to accommodate glass
      delivery  tubes.    The  exit  tube  must  be drawn  to  a  point  to  bubble
      diazomethane  through  the sample extract.   The generator assembly is shown
      in Figure 1.   The procedure for use  of this type of generator is given in
      Sec. 7.5.

      4.4   Other Glassware

            4.4.1 Beaker -  400 ml,  thick walled.

            4.4.2 Funnel -  75 mm  diameter.

            4.4.3 Separatory funnel  -  500 ml,  with Teflon stopcock.

            4.4.4 Centrifuge bottle -  500 ml (Pyrex 1260  or equivalent).

            4.4.5 Centrifuge bottle -  24/40 500 ml

            4.4.6 Continuous Extractor  (Hershberg-Wolfe type,  Lab Glass No. LG-
      6915, or equivalent)

            4.4.7 Pipet - Pasteur,  glass, disposable (140 mm x 5 mm  ID).

            4.4.8 Vials -  10 ml,  glass, with Teflon lined screw-caps.

            4.4.9 Volumetric flasks, Class A - 10 ml to  1000 ml.

      4.5   Filter  paper -  15 cm  diameter (Whatman No. 1  or equivalent).

      4.6   Glass Wool  - Pyrex, acid washed.

      4.7   Boiling    chips    -     Solvent    extracted    with    methylene
chloride,approximately 10/40 mesh (silicon carbide or equivalent).

      4.8   Water  bath  -  Heated,  with  concentric  ring   cover,  capable  of
temperature control (+ 2°C).   The bath  should  be used  in  a  hood.
                                   8151 - 5                         Revision 0
                                                                September 1994

-------
      4.9   Balance - Analytical, capable of accurately weighing to 0.0001 g.

      4.10  Centrifuge.

      4.11  Ultrasonic preparation - A horn-type device equipped with a titanium
tip, or a device that will give equivalent performance, shall be used.

            4.11.1      Ultrasonic Disrupter - The disrupter must have  a minimum
      power wattage of 300 watts, with pulsing capability.   A device designed to
      reduce the  cavitation  sound is  recommended.    Follow the manufacturers
      instructions for preparing the disrupter for extraction of samples.  Use
      a 3/4" horn for most samples.

      4.12  Sonabox -  Recommended with above disrupters for  decreasing cavitation
sound (Heat Systems - Ultrasonics, Inc., Model  432B or equivalent).

      4.13  pH paper.

      4.14  Silica gel cleanup  column (Bond Elut™ - Analytichem,  Harbor City, CA
            or equivalent).

5.0   REAGENTS

      5.1   Reagent grade inorganic chemicals shall  be  used in all tests.  Unless
otherwise  indicated,  it  is  intended  that all  reagents  shall  conform  to the
specifications of the Committee on Analytical Reagents of  the American Chemical
Society, where such  specifications  are  available.   Other  grades  may  be used,
provided it is  first ascertained that the reagent is  of sufficiently high purity
to permit its use without lessening the accuracy of the determination.

      5.2   Organic-free reagent water.   All  references to  water in this method
refer to organic-free water,  as defined in Chapter One.

      5.3   Sodium hydroxide  solution   (0.1  N),  NaOH.  Dissolve  4 g  NaOH in
organic-free reagent water and dilute to 1.0 L.

      5.4   Potassium  hydroxide  solution (37%  aqueous  solution  (w/v)),  KOH.
Dissolve 37  g  potassium  hydroxide pellets  in  organic-free  reagent water and
dilute to 100 mL.

      5.5   Phosphate buffer pH = 2.5 (0.1 M).   Dissolve 12 g sodium phosphate
(NaH2P04) in organic-free reagent water and dilute to 1.0 L.  Add  phosphoric acid
to adjust the pH to 2.5.

      5.6   N-methyl-N-nitroso-p-toluenesulfonamide (Diazald).   High  purity,
available from Aldrich Chemical Co.  or equivalent.

      5.7   Silicic acid, H2SiOs.   100 mesh powder,  store  at 130°C.

      5.8   Potassium carbonate, K2C03.

      5.9   2,3,4,5,6-Pentafluorobenzyl  bromide (PFBBr),  C6F5CH2Br.   Pesticide
quality or equivalent.


                                   8151  - 6                         Revision 0
                                                                September 1994

-------
      5.10  Sodium sulfate (granular, acidified, anhydrous), Na2S04.   Purify  by
heating at 400°C for 4  hours  in a shallow tray,  or by precleaning the  sodium
sulfate with  methylene chloride.   If  the  sodium  sulfate  is  precleaned with
methylene chloride, a method blank must  be analyzed, demonstrating that there  is
no interference  from the sodium sulfate.   Acidify by  slurrying 100 g  sodium
sulfate with enough  diethyl ether to  just  cover the solid; then add 0.1 ml  of
concentrated sulfuric acid and mix thoroughly.  Remove  the  ether under vacuum.
Mix  1  g  of the  resulting  solid with  5 ml of  organic-free reagent  water and
measure the pH of the mixture.   It must  be below a pH of 4.   Store the  remaining
solid at 130°C.
      5.11  Solvents

            5.11.1
      equivalent.

            5.11.2

            5.11.3

            5.11.4
Methylene  chloride,  CH2C12.    Pesticide  quality  or


Acetone, CH3COCH3.   Pesticide  quality or equivalent.

Methanol, CH3OH.  Pesticide quality or equivalent.

Toluene, C6H5CH3.  Pesticide quality  or  equivalent.
            5.11.5      Diethyl   Ether,   C2H5OC2H5.     Pesticide   quality  or
      equivalent.  Must  be  free of peroxides  as  indicated  by test strips  (EM
      Quant, or equivalent).  Procedures for removal of peroxides  are provided
      with the test strips.   After cleanup,  20  ml  of ethyl alcohol  preservative
      must be added to each liter of ether.
            5.11.6
      equivalent.

            5.11.7

            5.11.8
Isooctane,  (CH3)3CH2CH(CH3)2.     Pesticide  quality  or


Hexane, C6H14.   Pesticide quality or  equivalent.

Ethanol, absolute. C2H5OH
            5.11.9      Carbitol    (diethylene    glycol    monoethyl   ether),
      C2H5OCH2CH2OCH2CH20 -  optional  for producing alcohol-free diazomethane.

      5.12  Stock standard  solutions  (1000 tng/L) - Can  be  prepared from pure
standard materials or can be purchased as certified solutions.

            5.12.1      Prepare stock standard solutions  by accurately weighing
      about 0.010 g of  pure  acid.   Dissolve the material in pesticide quality
      acetone and dilute to volume in a 10 ml volumetric flask.  Stocks prepared
      from pure  methyl  esters  are  dissolved in  10%  acetone/isooctane  (v/v).
      Larger volumes may be used at the convenience of the analyst.   If compound
      purity is  certified  at 96% or greater,  the weight may  be  used without
      correction to calculate the concentration of the stock standard.

            5.12.2      Transfer the  stock  standard  solutions to  vials with
      Teflon lined  screw-caps.   Store at  4°C,  protected  from light.   Stock
      standard solutions should be checked frequently for signs of degradation
                                   8151 - 7
                                            Revision 0
                                        September 1994

-------
      or  evaporation,  especially  immediately  prior to  preparing calibration
      standards from them.

            5.12.3      Stock standard solutions of the derivatized acids must
      be replaced after  1 year,  or sooner,  if comparison with check standards
      indicates a problem.  Stock standard solutions of the free acids degrade
      more  quickly  and  should   be  replaced after  two  months,  or sooner  if
      comparison with check standards indicates a problem.

      5.13  Internal Standard Spiking Solution (if internal standard calibration
is used) - To  use this  approach,  the  analyst  must  select one or more internal
standards that are similar in analytical  behavior to the compounds of interest.
The  analyst  must further demonstrate  that  the measurement  of  the  internal
standard is not affected by  method or matrix interferences.  The compound 4,4'-
dibromooctafluorobiphenyl (DBOB)  has  been shown  to be  an  effective  internal
standard, but other  compounds, such as 1,4-dichlorobenzene, may be  used if there
is a DBOB interference.

            5.13.1      Prepare  an  internal   standard  spiking  solution  by
      accurately weighing approximately  0.0025  g  of pure DBOB.   Dissolve the
      DBOB  in  acetone  and   dilute  to  volume  in  a 10  ml volumetric  flask.
      Transfer the  internal   standard spiking solution  to  a  vial  with  a Teflon
      lined screw-cap,  and store at room  temperature.  Addition of 10 juL of the
      internal  standard spiking solution to 10 ml of sample extract results in
      a final  internal  standard concentration of 0.25 M9/L.  The solution should
      be replaced if there  is a change  in internal standard  response greater
      than 20 percent of the original  response recorded.

      5.14  Calibration standards - Calibration  standards, at  a minimum of five
concentrations  for  each  parameter of  interest,  should  be  prepared  through
dilution of  the stock  standards  with diethyl  ether  or  hexane.   One  of the
concentrations should be at a concentration near, but above, the method detection
limit.   The remaining concentrations should correspond to the expected range of
concentrations found in real samples or should define  the working range of the
GC.  Calibration  solutions   must  be  replaced  after six  months,  or sooner  if
comparison with check standards indicates a problem.

            5.14.1      Derivatize each  calibration  standard prepared from free
      acids  in a 10 ml  K-D  concentrator  tube, according  to the  procedures
      beginning at Sec. 7.5.

            5.14.2      Add   a known,  constant  amount  of one  or  more  internal
      standards to each derivatized calibration  standard, and dilute to volume
      with the solvent  indicated in the derivative  option used.

      5.15  Surrogate standards  - The analyst should monitor the performance of
the  extraction,  cleanup   (when   used),  and   determinative   step,   and   the
effectiveness of the method  in dealing with each sample matrix, by spiking  each
sample,   standard,   and  blank with one  or two herbicide  surrogates  (e.g.,
herbicides that are not  expected to be present in  the sample)  recommended  to
encompass the range  of  the temperature program used  in this method.  Deuterated
analogs of  analytes  should  not  be  used  as  surrogates in gas chromatographic
                                   8151  - 8                         Revision 0
                                                                September 1994

-------
analysis due to coelution problems.  The  surrogate standard recommended for use
is 2,4-Dichlorophenylacetic acid (DCAA).

            5.15.1      Prepare  a  surrogate  standard  spiking  solution  by
      accurately weighing approximately 0.001 g of pure DCAA.  Dissolve the DCAA
      in acetone, and dilute to volume in a 10 ml volumetric flask.   Transfer
      the surrogate  standard  spiking  solution to a  vial  with  a  Teflon  lined
      screw-cap,  and  store at  room  temperature.   Addition  of  50 juL of the
      surrogate standard spiking solution to 1 L of sample, prior to extraction,
      results in a final concentration in the extract of 0.5 mg/L.

      5.16  pH Adjustment Solutions

            5.16.1      Sodium hydroxide, NaOH, 6 N.

            5.16.2      Sulfuric acid, H2S04, 12  N.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See the  introductory  material  to this chapter,  Organic  Analytes,
Sec.  4.1.  1 L samples should be collected.

      6.2   Extracts must be stored under refrigeration (4°C).


7.0   PROCEDURE

      7.1   Preparation of High Concentration Waste  Samples

            7.1.1 Extraction

                  7.1.1.1     Follow  Method  3580,  Waste  Dilution,  with  the
            following exceptions:

                  •     use diethyl  ether as  the  dilution  solvent,
                  •     use acidified  anhydrous  sodium  sulfate,  and  acidified
                        glass wool,
                  •     spike the sample with  surrogate compound(s) according to
                        Sec. 5.16.1.

                  7.1.1.2     If the sample is to be analyzed  for both herbicide
            esters and acids, then the sample extract must be  hydrolyzed.   In
            this case, transfer 1.0 mL (a smaller volume or  a  dilution may be
            required if herbicide concentrations  are  large)  to  a 250  mL ground
            glass Erlenmeyer flask.   Proceed  to Sec.  7.2.1.8.  If the analysis
            is  for   acid   herbicides  only,  proceed  to   Sec.  7.4.5   for
            derivatization by diazomethane (if  PFB derivatization is  selected,
            reduce the volume  of diethyl ether to 0.1  - 0.5 mL as per Sec.  7.4.2
            and then dilute to 4 mL  with  acetone).
                                   8151  -  9                         Revision  0
                                                                September 1994

-------
7.2   Preparation of Soil,  Sediment, and Other Solid Samples

      7.2.1 Extraction

            7.2.1.1     To  a  400 ml,  thick-wall  beaker  add  30  g  (dry
      weight as determined  in Method 3540, Sec. 7.2.1) of the well-mixed
      solid sample.   Adjust  the pH to  2  with  concentrated hydrochloric
      acid or acidify solids  in  each beaker with 85 ml of 0.1 M phosphate
      buffer  (pH  = 2.5) and  thoroughly mix the  contents with  a glass
      stirring rod.  Spike the sample with  surrogate compound(s) according
      to Sec. 5.16.1.

            7.2.1.2     The  ultrasonic extraction  of  solids  must  be
      optimized for  each type of sample.    In  order  for  the  ultrasonic
      extractor to efficiently extract solid  samples,  the sample must be
      free flowing when  the solvent  is  added.  Acidified anhydrous sodium
      sulfate should be added to clay  type soils  (normally 1:1),  or any
      other solid that is not a free flowing  sandy mixture, until  a free
      flowing mixture is obtained.

            7.2.1.3     Add 100  ml  of  methylene  chloride/acetone  (1:1
      v/v) to the  beaker.  Perform  ultrasonic extraction  for  3  minutes,
      with output control knob set at 10 (full power) and with mode switch
      on Pulse (pulsing  energy rather than  continuous energy) and percent-
      duty cycle knob set at  50% (energy  on  50% of time  and  off  50% of
      time).   Allow the  solids to settle.  Transfer the organic layer into
      a 500 mL centrifuge bottle.

            7.2.1.4     Ultrasonically extract the sample twice more using
      100 ml of methylene chloride  and the same ultrasonic conditions.

            7.2.1.5     Combine the  three organic extracts  from the sample
      in the centrifuge bottle  and  centrifuge  10 minutes  to  settle the
      fine particles.   Filter the combined extract  through filter paper
      (Whatman #1, or equivalent) containing 7-10 g of  acidified  sodium
      sulfate into a 500 ml  24/40 Erlenmeyer flask.  Add 10 g of acidified
      anhydrous  sodium  sulfate.    Periodically,  vigorously  shake  the
      extract and drying agent  and  allow  the drying agent to remain  in
      contact with the extract for a minimum of  2 hours.  See NOTE in Sec.
      7.3.1.6  that  emphasizes  the need  for  a  dry  extract  prior  to
      esterification.

            7.2.1.6     Quantitatively transfer the contents of the flask
      to a  500-mL  Kuderna-Danish fTask with  a 10-mL concentrator  tube
      attached.   Add  boiling  chips  and attach the macro  Snyder  column.
      Evaporate the extract on the water bath  to a volume of approximately
      5 ml.   Remove  the flasks from the water  bath and allow them to cool.

            7.2.1.7     If  hydrolysis  or  additional   cleanup   is   not
      required and  the  sample  is dry,  proceed  to Sec. 7.4.4  -  Nitrogen
      Slowdown.
                            8151 - 10                         Revision  0
                                                          September 1994

-------
      7.2.1.8     Use this step only if herbicide esters  in addition
to herbicide acids are to be determined:

            7.2.1.8.1   Add 5 ml  of 37%  aqueous potassium hydroxide
      and 30 ml  of water to the extract.   Add  additional  boiling
      chips to the  flask.   Reflux the  mixture  on  a  water  bath at
      60-65°C  until  the hydrolysis  step  is completed (usually 1 to
      2 hours).  Remove the flasks from the water bath and cool to
      room temperature. CAUTION - the presence of residual  acetone
      will result  in the  formation of aldol  condensation products
      which will  cause GC interference.

            7.2.1.8.2   Transfer the  hydrolyzed  aqueous solution to
      a 500 ml separatory funnel  and extract the  solution  three
      times with  100 ml portions of methylene chloride.  Discard the
      methylene chloride phase.   At this point the basic (aqueous)
      solution contains the herbicide salts.

            7.2.1.8.3   Adjust the pH of the solution  to  <2 with
      cold (4°C)  sulfuric  acid (1:3) and extract once with 40 ml of
      diethyl  ether and twice with 20  ml  portions  of ether.  Combine
      the extracts and  pour them through a pre-rinsed drying column
      containing  7 to 10 cm of acidified anhydrous sodium sulfate.
      Collect  the dried extracts  in a  500 ml Erlenmeyer flask (with
      a 24/40  joint)  containing  10  g  of  acidified anhydrous sodium
      sulfate.    Periodically, vigorously shake  the extract  and
      drying agent and allow the  drying  agent to remain  in  contact
      with the extract for a minimum  of  2 hours.  See NOTE  in Sec.
      7.3.1.6  that emphasizes the need  for  a  dry extract prior to
      esterification.   Quantitatively transfer the contents of the
      flask  to  a  500-mL  Kuderna-Danish   flask  with   a  10-mL
      concentrator tube attached when the extract is known  to be
      dry.

            7.2.1.8.4   Proceed to Sec.  7.4, Extract Concentration.
      If additional cleanup is required, proceed to Sec.  7.2.1.9.

      7.2.1.9      Use this step if additional cleanup of the non-
hydrolyzed herbicides is required:

            7.2.1.9.1    Partition the herbicides by extracting the
      methylene  chloride   from  7.2.1.7   (or  diethyl  ether  from
      7.2.1.8.4)   with  3 x 15 ml  portions of aqueous  base prepared
      by carefully mixing  30  ml  of  reagent water into 15 ml of 37%
      aqueous  potassium hydroxide.  Discard the  methylene chloride
      or ether phase.   At this point  the  basic  (aqueous)  solution
      contains  the herbicide  salts.

            7.2.1.9.2    Adjust the pH of  the solution to  <2 with
      cold (4°C)  sulfuric  acid (1:3) and extract once with 40 ml of
      diethyl  ether and twice with 20  ml  portions  of ether.  Combine
      the extracts and  pour them through a pre-rinsed drying column
      containing  7 to  10 cm of acidified anhydrous  sodium sulfate.
                      8151 - 11                         Revision 0
                                                    September 1994

-------
            Collect the dried extracts  in  a  500 ml Erlenmeyer flask (with
            a 24/40 joint) containing 10 g of acidified anhydrous sodium
            sulfate.   Periodically,  vigorously  shake  the extract  and
            drying agent and allow the drying agent to remain in contact
            with the extract for a minimum of 2 hours.  See NOTE in Sec.
            7.3.1.6 that emphasizes the need  for a  dry extract prior to
            esterification.   Quantitatively transfer the contents of the
            flask  to  a  500-mL  Kuderna-Danish   flask   with   a  10-mL
            concentrator tube attached  when the  extract  is known  to be
            dry.

                  7.2.1.9.3    Proceed   to  section   7.4   for   extract
            concentration.

            7.2.1.10    An alternative wrist-shaker extraction procedure
      can be found in Sec. 7.2 of Method 8150.

7.3   Preparation of Aqueous Samples

      7.3.1 Separatory Funnel

            7.3.1.1     Using a  graduated cylinder,  measure  out  a  1-L
      sample and  transfer  it into a 2-L  separatory funnel.    Spike  the
      sample with surrogate  compound(s)  according  to Sec.  5.15.1.

            7.3.1.2     Add  250  g of NaCl  to the  sample,  seal,  and shake
      to dissolve the salt.

            7.3.1.3     Use  this  step only if herbicide esters in addition
      to herbicide acids,  are to be  determined:

                  7.3.1.3.1    Add 17 mL  of 6 N NaOH to the sample,  seal,
            and shake.  Check the pH of  the sample  with pH paper; if the
            sample does not  have  a pH greater  than or equal to 12, adjust
            the pH by adding more 6 N NaOH.   Let the  sample  sit at room
            temperature until the hydrolysis step is completed (usually 1
            to  2  hours),   shaking  the  separatory   funnel  and  contents
            periodically.

                  7.3.1.3.2    Add 60  mL  of methylene  chloride to  the
            sample  bottle and  rinse both  the bottle   and  the  graduated
            cylinder. Transfer the methylene  chloride  to  the  separatory
            funnel and extract the sample by vigorously  shaking the funnel
            for  2  minutes,  with periodic  venting  to  release  excess
            pressure.  Allow the  organic layer to separate from the water
            phase for a minimum of 10 minutes.  If the  emulsion interface
            between the layers is more  than one-third  the volume of  the
            solvent layer,  the analyst must employ  mechanical  techniques
            to  complete the phase  separation.   The   optimum  technique
            depends upon the sample, but may include stirring,  filtration
            through glass wool, centrifugation, or other physical methods.
            Discard the methylene chloride phase.
                            8151 - 12                         Revision  0
                                                          September 1994

-------
            7.3.1.3.3   Add  a  second  60  mL  volume  of methylene
      chloride to  the  separatory funnel  and repeat the extraction
      procedure  a  second  time,  discarding  the  methylene chloride
      layer.  Perform a third extraction in the same manner.

      7.3.1.4     Add 17 mL of cold (4°C) 12 N sulfuric  acid to the
sample (or hydrolyzed sample),  seal, and shake to mix.  Check the pH
of the sample with pH paper:   if the sample does not have a pH less
than or equal to 2, adjust the pH by adding more acid.

      7.3.1.5     Add 120 mL diethyl ether  to the sample, seal, and
extract the sample by vigorously shaking the funnel for 2 min with
periodic  venting to release  excess pressure.   Allow  the  organic
layer to separate from the water phase  for  a minimum of  10 min.  If
the emulsion  interface  between layers  is  more  than  one third the
volume of  the solvent  layer,  the analyst must  employ  mechanical
techniques to complete  the phase separation. The optimum techniques
to complete the  phase  separation  depends upon  the  sample,  but may
include stirring, filtration through glass  wool, centrifugation, or
other  physical   methods.    Remove  the aqueous  phase  to   a  2  L
Erlenmeyer flask and collect the ether  phase in a 500 mL Erlenmeyer
flask containing approximately  10 g of acidified  anhydrous sodium
sulfate.    Periodically, vigorously shake  the  extract  and  drying
agent.

      7.3.1.6     Return the aqueous phase to the separatory funnel,
add 60 mL of diethyl  ether  to  the  sample, and repeat the extraction
procedure  a second  time,  combining the  extracts  in  the  500  mL
Erlenmeyer  flask.   Perform a third extraction with  60  mL  diethyl
ether in the  same  manner.   Allow the extract  to remain  in  contact
with the sodium sulfate for approximately 2 hours.

      NOTE:       The  drying  step  is  very  critical  to  ensuring
                  complete esterification.   Any moisture remaining
                  in  the  ether  will   result   in  low  herbicide
                  recoveries.   The amount of  sodium  sulfate  is
                  adequate  if  some  free   flowing  crystals  are
                  visible when swirling the flask.   If  all  of the
                  sodium sulfate  solidifies  in  a cake,   add  a few
                  additional grams of  acidified sodium sulfate and
                  again test by  swirling.  The  2  hour drying time
                  is a  minimum,  however, the extracts may  be held
                  in contact with the  sodium sulfate overnight.

      7.3.1.7     Pour  the dried  extract through  a funnel  plugged
with acid  washed glass wool, and collect  the  extract  in  the K-D
concentrator.  Use  a  glass  rod to crush any caked  sodium  sulfate
during the transfer.   Rinse  the  Erlenmeyer  flask and funnel  with 20
to 30 mL  of diethyl  ether to complete  the quantitative transfer.
Proceed to Sec.  7.4 for extract concentration.
                      8151  -  13                         Revision 0
                                                    September 1994

-------
7.4   Extract Concentration

      7.4.1 Add one or two clean boiling chips to the flask and attach a
three ball Snyder column.  Prewet the Snyder column by adding about 1 ml
of diethyl ether to the top of the column.  Place the K-D apparatus on a
hot water bath  (15-20°C above the boiling point of the solvent)  so that the
concentrator tube is partially  immersed  in  the  hot  water and the entire
lower rounded surface of the flask is bathed with hot vapor.  Adjust the
vertical position of the apparatus and the water temperature, as required,
to complete the  concentration  in 10-20 minutes.   At  the  proper  rate of
distillation the balls  of the  column  will  actively chatter,  but  the
chambers will not flood.  When the apparent volume of liquid  reaches 1 ml,
remove the K-D  apparatus  from  the water bath and allow  it  to drain and
cool for at least 10 minutes.

      7.4.2 Remove  the Snyder  column and rinse the  flask and its lower
joints  into  the concentrator  tube  with  1-2 ml  of  diethyl ether.   The
extract  may  be  further  concentrated by  using  either the  micro  Snyder
column technique (Sec. 7.4.3) or nitrogen  blowdown  technique  (Sec. 7.4.4).
      7.4.3 Micro Snyder Column Technique

            7.4.3.1     Add another one or two  clean boiling chips to the
      concentrator tube and attach a two ball  micro Snyder column.  Prewet
      the column by adding  about 0.5 ml of diethyl  ether to the top of the
      column.   Place  the K-D  apparatus in a hot  water  bath  so  that the
      concentrator tube is  partially  immersed  in  the  hot  water.   Adjust
      the vertical position of the apparatus and the water temperature, as
      required, to  complete the  concentration  in  5-10  minutes.   At the
      proper rate of  distillation the  balls  of the column  will  actively
      chatter,  but the chambers will not  flood.  When the apparent volume
      of liquid reaches 0.5 ml, remove the K-D apparatus  from the water
      bath and  allow it to  drain and cool  for  at least  10 minutes.  Remove
      the Snyder  column and rinse the  flask and  its  lower  joints  with
      about 0.2 ml  of diethyl  ether  and add to  the  concentrator tube.
      Proceed to Sec. 7.4.5.

      7.4.4 Nitrogen Blowdown Technique

            7.4.4.1     Place the concentrator tube in a warm water bath
      (approximately  35°C)  and  evaporate  the  solvent volume  to  the
      required  level   using a  gentle  stream   of clean,  dry  nitrogen
      (filtered through a column of activated carbon).

            CAUTION:    Do  not use plasticized  tubing between the carbon
                        trap and the sample.

            7.4.4.2     The internal wall  of  the tube must be rinsed down
      several  times  with  diethyl  ether  during  the  operation.    During
      evaporation, the  solvent  level  in  the tube  must  be  positioned to
      prevent  water  from condensing  into the  sample  (i.e.,  the  solvent
      level should be  below the level  of the water bath).   Under normal
                             8151  -  14                         Revision 0
                                                          September 1994

-------
            operating conditions, the extract  should  not be allowed to become
            dry.  Proceed to Sec. 7.4.5.

            7.4.5 Dilute  the  extract with  1  ml  of  isooctane  and 0.5  mL  of
      methanol.    Dilute  to a final  volume  of 4  mL with diethyl  ether.   The
      sample is  now ready for  methylation  with  diazomethane.  If PFB derivation
      is being performed, dilute to 4 ml with acetone.

      7.5   Esterification - For diazomethane derivatization proceed with Sec.
7.5.1.  For PFB  derivatization proceed with  Sec.  7.5.2.

            7.5.1 Diazomethane Derivatization - Two  methods  may be  used for the
      generation  of  diazomethane:    the bubbler method  (see Figure  1),  Sec.
      7.5.1.1, and the Diazald kit method, Sec. 7.5.1.2.
            CAUTION:    Diazomethane  is  a  carcinogen
                        certain conditions.
                         and  can  explode  under
            The  bubbler  method  is  suggested  when  small  batches of  samples
      (10-15)  require  esterification.   The  bubbler  method  works well  with
      samples that  have low concentrations of herbicides  (e.g., aqueous samples)
      and is  safer to use than  the Diazald  kit  procedure.   The  Diazald  kit
      method is  good  for  large quantities of  samples  needing esterification.
      The Diazald kit method is more effective than the bubbler method for soils
      or  samples that may contain  high  concentrations of  herbicides  (e.g.,
      samples  such as  soils   that  may  result  in  yellow extracts  following
      hydrolysis may  be  difficult  to  handle by the   bubbler  method).    The
      diazomethane  derivatization (U.S.EPA,  1971)  procedures, described below,
      will react efficiently with all of the  chlorinated  herbicides described in
      this method and should be  used  only by  experienced  analysts,  due to  the
      potential hazards associated  with its  use.   The  following  precautions
      should be taken:

                 Use a safety  screen.
                 Use mechanical  pipetting aides.
                 Do not heat  above  90°C  - EXPLOSION may result.
                 Avoid grinding surfaces, ground-glass joints, sleeve bearings,
                 and glass  stirrers - EXPLOSION may result.
                 Store away  from alkali  metals -  EXPLOSION may  result.
                 Solutions of  diazomethane decompose rapidly in the presence of
                 solid materials such as copper powder, calcium chloride,  and
                 boiling  chips.
                  7.5.1.1
            (see Figure 1).
Bubbler method - Assemble the diazomethane bubbler
                        7.5.1.1.1    Add  5 mL of diethyl ether to the first test
                  tube.   Add  1 mL of diethyl ether, 1 mL of carbitol,  1.5 mL of
                  37% KOH,  and 0.1-0.2  g of  Diazald  to the second test  tube.
                  Immediately place the exit  tube into the concentrator  tube
                  containing   the  sample  extract.     Apply  nitrogen   flow
                  (10 mL/min) to bubble diazomethane  through  the  extract  for
                  10 minutes  or until the yellow color of diazomethane  persists.
                                  8151 - 15
                                      Revision 0
                                  September 1994

-------
      The amount of Diazald  used is sufficient for esterification of
      approximately three sample extracts.   An  additional 0.1-0.2 g
      of  Diazald  may  be   added  (after  the  initial   Diazald  is
      consumed) to extend the generation of the diazomethane.  There
      is sufficient KOH present in the original  solution to perform
      a maximum of approximately 20 minutes of total esterification.

            7.5.1.1.2   Remove  the  concentrator  tube  and  seal  it
      with a Neoprene or Teflon stopper.  Store  at room  temperature
      in a hood for 20 minutes.

            7.5.1.1.3   Destroy any unreacted di azomethane by adding
      0.1-0.2 g of silicic  acid to the concentrator tube.  Allow to
      stand until  the  evolution of nitrogen gas has stopped.  Adjust
      the  sample   volume  to  10.0  mL  with hexane.    Stopper  the
      concentrator tube or  transfer 1  ml of sample to  a  GC vial, and
      store refrigerated if  further processing will not  be performed
      immediately.  Analyze by gas chromatography.

            7.5.1.1.4   Extracts should be  stored at 4°C away from
      light.  Preservation study results indicate that most analytes
      are stable  for  28  days;  however,  it  is recommended that the
      methylated  extracts be  analyzed immediately to minimize the
      trans-esterification  and  other  potential  reactions that may
      occur.

      7.5.1.2      Diazald kit  method  -  Instructions  for preparing
diazomethane are  provided with the generator kit.

            7.5.1.2.1   Add 2  mL  of  diazomethane solution and let
      the sample stand for 10 minutes with occasional  swirling.  The
      yellow  color of diazomethane  should be  evident  and  should
      persist for  this period.

            7.5.1.2.2   Rinse the inside wall of the ampule with 700
      /iL  of   diethyl  ether.     Reduce   the   sample   volume  to
      approximately 2 ml to remove excess diazomethane  by allowing
      the solvent  to  evaporate spontaneously  at room temperature.
      Alternatively,  10 mg of silicic acid can be added to destroy
      the excess diazomethane.

            7.5.1.2.3   Dilute the sample  to  10.0 ml with hexane.
      Analyze by  gas  chromatography.   It  is  recommended  that the
      methylated  extracts be  analyzed immediately to minimize the
      trans-esterification and other  potential  reactions  that may
      occur.

7.5.2 PFB Method

      7.5.2.1     Add 30 /zL of 10% K2C03  and 200 /zL  of 3% PFBBr in
acetone.  Close the tube with  a  glass stopper  and mix  on  a  vortex
mixer.  Heat the tube at 60°C  for 3 hours.
                       8151  -  16                         Revision 0
                                                    September 1994

-------
            7.5.2.2     Evaporate the  solution  to 0.5 ml  with  a gentle
      stream of nitrogen.   Add 2  mL  of  hexane and  repeat evaporation just
      to dryness at ambient temperature.

            7.5.2.3     Redissolve the residue in 2 ml of toluene:hexane
      (1:6) for column cleanup.

            7.5.2.4     Top a  silica  column (Bond  Elut™  or equivalent)
      with 0.5 cm of anhydrous sodium sulfate.   Prewet the column with 5
      ml hexane and  let  the solvent drain to the top of the adsorbent.
      Quantitatively transfer  the reaction residue  to the  column  with
      several rinsings of the toluene:hexane solution (total 2-3 mL).

            7.5.2.5     Elute the column with sufficient toluene:hexane to
      collect  8  ml of  eluent.    Discard  this  fraction,  which  contains
      excess reagent.

            7.5.2.6     Elute  the column  with  toluene:hexane  (9:1)  to
      collect  8  ml  of  eluent  containing PFB  derivatives  in  a 10  mL
      volumetric flask.   Dilute to 10 mL with hexane.  Analyze by GC/ECD.

7.6   Gas chromatographic conditions (recommended):

      7.6.1 Narrow Bore

            7.6.1.1     Primary Column 1:

            Temperature program:     60°C  to 300°C,  at  4°C/min
            Helium carrier flow:     30 cm/sec
            Injection volume:        2 /iL,  splitless,  45 sec delay
            Injector temperature:   250°C
            Detector temperature:   320°C

            7.6.1.2     Primary Column la:

            Temperature program:     60°C  to 300°C,  at  4°C/min
            Helium carrier flow:     30 cm/sec
            Injection volume:        2 juL,  splitless,  45 sec delay
            Injector temperature:   250°C
            Detector temperature:   320°C

            7.6.1.3     Column 2:

            Temperature program:     60°C  to 300°C,  at  4°C/min
            Helium carrier flow:     30 cm/sec
            Injection volume:        2 juL,  splitless,  45 sec delay
            Injector temperature:   250°C
            Detector temperature:   320°C
                            8151  -  17                         Revision 0
                                                          September 1994

-------
            7.6.1.4     Confirmation Column:

            Temperature program:    60°C to 300°C,  at 4°C/min
            Helium carrier flow:    30  cm/sec
            Injection volume:       2 jiiL,  splitless,  45  sec delay
            Injector temperature:   250°C
            Detector temperature:   320°C

      7.6.2 Wide-bore

            7.6.2.1     Primary Column:

            Temperature program:    0.5 minute at 150°C,  150°C to 270°C  at
                                    5°C/min
            Helium carrier flow:    7 mL/min
            Injection volume:       1 /xL

            7.6.2.2     Confirmatory Column:

            Temperature program:    0.5 minute at 150°C,  150°C to 270°C  at
                                    5°C/min
            Helium carrier flow:    7 mL/min
            Injection volume:       1 p,l

7.7   Calibration

      7.7.1 The  procedure  for  internal  or external  calibration  may  be
used.  Refer to Method 8000 for a description of each of these procedures.
Use Table 1 for  guidance on  selecting the  lowest  point on the calibration
curve.

7.8   Gas chromatographic analysis

      7.8.1 Refer to Method  8000.   If  the internal standard calibration
technique is used, add 10 /zL of internal standard  to  the sample prior  to
injection.

      7.8.2 Follow Method 8000 for instructions on  the analysis sequence,
appropriate dilutions,  establishing daily  retention time windows,  and
identification criteria.  Include  a mid-concentration  standard after each
group of 10 samples in the analysis sequence.

      7.8.3 An example of a  chromatogram  for a methylated chlorophenoxy
herbicide is shown in Figure  2.   Tables 2 and 3 present retention times
for  the  target  analytes  after esterification,  using  the  diazomethane
derivatization   procedure    and   the   PFB   derivatization   procedure,
respectively.

      7.8.4 Record the sample volume injected and the resulting peak sizes
(in area units or peak heights).

      7.8.5 Using either the  internal  or  external  calibration procedure
(Method 8000),  determine the  identity and  quantity  of  each component peak


                             8151  -  18                         Revision 0
                                                          September 1994

-------
      in the  sample  chromatogram which corresponds to  the  compounds  used for
      calibration purposes.

            7.8.6 If calibration standards have been analyzed in the same manner
      as the samples (e.g. have undergone hydrolysis and esterification), then
      the calculation  of concentration given  in  Method 8000  should  be used.
      However,  if  calibration  is performed  using standards made  from methyl
      ester compounds (compounds not  esterified by application of this method),
      then the calculation of concentration  must  include  a  correction for the
      molecular weight of the methyl  ester versus the acid herbicide.

            7.8.7 If peak  detection  and  identification  are prevented  due  to
      interferences, further cleanup is required.   Before using  any cleanup
      procedure, the  analyst  must process  a series of standards  through the
      procedure to validate elution  patterns and  the  absence  of interferences
      from reagents.


8.0   QUALITY CONTROL

      8.1   Refer  to Chapter  One for  specific  quality  control  procedures.
Quality control to validate sample extraction is covered in Method 3500 and in
the extraction method utilized.   If extract cleanup was performed, follow the QC
in Method 3600 and in the specific cleanup method.

      8.2   Procedures to check the GC system operation are found in Method 8000.

            8.2.1 Select a representative  spike concentration for each compound
      (acid or ester) to be measured.  Using  stock standards, prepare a quality
      control  check  sample concentrate, in  acetone,  that  is  1000  times  more
      concentrated than the selected  concentrations.  Use this quality control
      check sample concentrate  to prepare quality control  check samples.

            8.2.2 Tables 4 and  5 present bias and precision data for water and
      clay matrices, using the  diazomethane derivatization procedure.   Table 6
      presents relative  recovery data generated using  the  PFB derivatization
      procedure and water samples. Compare the results obtained with the results
      given in these Tables to  determine if the data quality is acceptable.

      8.3   Calculate surrogate  standard  recovery on all  standards,  samples,
blanks,  and   spikes.    Determine if  the  recovery  is  within   limits  (limits
established by performing QC  procedures outlined in Method 8000).

            8.3.1 If recovery is  not within limits,  the following procedures are
      required:

                  8.3.1.1     Check  to  be  sure   there  are   no   errors  in
            calculations,  surrogate  solutions  and internal  standards.   Also,
            check instrument  performance.

                  8.3.1.2     Recalculate the data and/or reanalyze the extract
            if any of the above checks reveal a problem.
                                  8151  -  19                         Revision 0
                                                                September 1994

-------
                  8.3.1.3     Reextract and reanalyze the sample  if none of the
            above are a problem or flag the data as "estimated concentration."

      8.4   GC/MS confirmation

            8.4.1 GC/MS techniques  should  be judiciously  employed  to support
      qualitative identifications made with this method.  Refer to Method 8270
      for the appropriate GC/MS operating conditions and analysis procedures.

            8.4.2 When  available,   chemical  ionization  mass  spectra may  be
      employed to aid the qualitative identification process.

            8.4.3 Should  these MS  procedures  fail  to provide  satisfactory
      results, additional  steps may  be taken before reanalysis.  These  steps may
      include the use of alternate packed or capillary GC columns or additional
      cleanup.


9.0   METHOD PERFORMANCE

      9.1   In single laboratory studies using  organic-free  reagent  water and
clay/still bottom samples, the mean  recoveries presented in Tables 4 and 5 were
obtained for diazomethane derivatization. The standard deviations  of the percent
recoveries of these measurements are also in Tables 4 and 5.

      9.2   Table 6  presents relative recoveries of the target analytes obtained
using the PFB derivatization procedure with spiked water samples.


10.0  REFERENCES

1.    Fed. Reg.  1971, 38,  No.  75,  Pt.  II.

2.    Goerlitz,  D. G.; Lamar, W.L., "Determination of Phenoxy Acid Herbicides in
      Water by Electron  Capture and Microcoulometric Gas  Chromatography,".  U.S.
      Geol. Survey Water Supply Paper 1967, 1817-C.

3.    Burke, J.  A.  "Gas  Chromatography  for  Pesticide  Residue Analysis;  Some
      Practical  Aspects, J. Assoc.  Off Anal.  Chem. 1965, 48,  1037.

4.    "Extraction and Cleanup  Procedures for the Determination of Phenoxy Acid
      Herbicides  in Sediment";  U.S.  Environmental  Protection  Agency.    EPA
      Toxicant and Analysis Center:  Bay St. Louis, MS,  1972.

5.    Shore, F.L.; Amick,  E.N.;  Pan, S. T.  "Single  Laboratory Validation of EPA
      Method  8151  for  the Analysis  of Chlorinated  Herbicides  in  Hazardous
      Waste";  U.S.  Environmental Protection Agency.   Environmental  Monitoring
      Systems  Laboratory.  Office of Research  and  Development, Las  Vegas,  NV,
      1985; EPA-60014-85-060.

6.    Method  515.1,  "Determination  of  Chlorinated  Acids  in  Water by  Gas
      Chromatography with  an  Electron Capture Detector",  Revision  4.0,  USEPA,
      Office  of   Research  and Development,  Environmental  Monitoring Systems
      Laboratory, Cincinnati,  Ohio.

                                  8151  - 20                         Revision 0
                                                                September 1994

-------
7.    Method 1618,  "Organo-halide and Organo-phosphorus  Pesticides and Phenoxy-
      acid Herbicides  by Wide Bore  Capillary Column Gas  Chromatography with
      Selective  Detectors",  Revision  A,  July  1989, USEPA,  Office  of  Water
      Regulations and Standards,  Washington, DC.

8.    Gurka, D.F, Shore, F.L., Pan,  S-T,  "Single Laboratory Validation of EPA
      Method 8150  for  Determination  of  Chlorinated Herbicides  in Hazardous
      Waste", JAOAC,  69, 970, 1986.
                                   8151  -  21                         Revision 0
                                                                September 1994

-------
                                Figure 1
                         DIAZOMETHANE GENERATOR
    nitrogen
rubber  (topper
                                                                  gloss tubing

                     tube 1
tube 2
                                  8151  - 22
                        Revision  0
                    September  1994

-------
                                            Figure 2
                  CHROMATOGRAM OF  METHYL ESTERS  OF  CHLOROPHENOXYACIDS
100.0
 RIC-
    200             400
    3:2O             6:40
 600
10:OO
                                                                                Dalapon. methyl ester
                                                                                Dicamba. methyl ester
                                                                                MCPP. methyl estsr
                                                                                MCPA. methyl ester
                                                                                Dichlorprop. methyl ester
                                                                                2.4.-D. methyl ester
                                                                                Silver, nieihyl ester
                                                                                2.4.6 T. methyl ester
                                                                                2.4 OB. methyl ester
                                                                                Oinoseli. methyl ether
1200
20:00
                                         Scan Time
                                            8151  - 23
                                                  Revision  0
                                             September 1994

-------
                                    TABLE 1
              ESTIMATED METHOD DETECTION LIMITS FOR METHOD 8151,
                          DIAZOMETHANE  DERIVATIZATION
Aqueous Samples




Analyte
Acifluorfen
Bentazon
Chloramben
2,4-0
Dalapon
2,4-DB
DCPA diacid"
Dicamba
3,5-Dichlorobenzoic acid
Dichloroprop
Dinoseb
5-Hydroxydicamba
MCPP
MCPA
4-Nitrophenol
Pentachlorophenol
Picloram
2,4,5-T
2,4,5-TP
GC/ECD
Estimated
Detection
Limit8
(M9/L)
0.096
0.2
0.093
0.2
1.3
0.8
0.02
0.081
0.061
0.26
0.19
0.04
0.09d
0.056d
0.13
0.076
0.14
0.08
0.075
Soil Samples
GC/ECD
Estimated
Detection
Limitb
(/*g/kg)


4.0
0.11
0.12



0.38



66
43
0.34
0.16


0.28
GC/MS
Estimated
Identification
Limit0
(ng)


1.7
1.25
0.5



0.65



0.43
0.3
0.44
1.3


4.5
a  EDL = estimated detection limit;  defined  as  either  the MDL  (40 CFR Part 136,
   Appendix B,  Revision 1.11  ),  or  a  concentration of  analyte  in  a  sample
   yielding  a  peak  in the   final   extract  with  signal-to-noise  ratio  of
   approximately 5, whichever value is higher.

b  Detection limits determined from standard solutions  corrected  back to 50 g
   samples,  extracted  and  concentrated  to  10  mL,  with  5  /xL  injected.
   Chromatography   using   narrow  bore  capillary   column,   0.25  ^m   film,
   5% pnenyl/95% methyl silicone.

c  The minimum amount of analyte to give a  Finnigan  INCOS  FIT value of  800 as
   the methyl  derivative vs. the  spectrum obtained from  50 ng of the respective
   free acid herbicide.
   40 CFR Part 136, Appendix B  (49  FR  43234).
   capillary column.
Chromatography using wide-bore
e  DCPA monoacid and diacid metabolites  included  in  method  scope;  DCPA diacid
   metabolite used for validation studies.  DCPA is a dimethyl  ester.
                                  8151  -  24
                    Revision 0
                September 1994

-------
                                  TABLE 2
RETENTION TIMES (MINUTES) OF METHYL DERIVATIVES OF  CHLORINATED  HERBICIDES
 Megabore Columns
                        Narrow Bore Columns
                     Wide-bore  Columns
Analyte
Dalapon
3,5-Dichlorobenzoic
4-Nitrophenol
DCAA (surrogate)
Dicamba
Dichloroprop
2,4-D
DBOB (internal std.
Pentachlorophenol
Chloramben
2,4,5-TP
5-Hydroxydicamba
2,4,5-T
2,4-DB
Dinoseb
Bentazon
Picloram
DCPA diacidc
Acifluorfen
MCPP
MCPA
Primary8
Column
3.4
acid 18.6
18.6
22.0
22.1
25.0
25.5
) 27.5
28.3
29.7
29.7
30.0
30.5
32.2
32.4
33.3
34.4
35.8
41.5


Confirmation8
Column
4.7
17.7
20.5
14.9
22.6
25.6
27.0
27.6
27.0
32.8
29.5
30.7
30.9
32.2
34.1
34.6
37.5
37.8
42.8


Primary6
Column




4.39
5.15
5.85



6.97

7.92
8.74





4.24
4.74
Confirmation5
Column




4.39
5.46
6.05



7.37

8.20
9.02





4.55
4.94
 Primary Column:
 Confirmation Column:
          Temperature program:
          Helium carrier flow:
          Injection volume:
          Injector temperature:
          Detector temperature;
 Primary Column:
 Confirmatory Column:
          Temperature program:

          Helium carrier flow:
          Injection volume:
5% phenyl/95% methyl silicone
14% cyanopropyl phenyl silicone
60°C  to 300°C,  at  4°C/min
30 cm/sec
2 juL,  splitless,  45 sec  delay
250°C
320°C
DB-608
14% cyanopropyl phenyl silicone
0.5 minute at 150°C,
150°C to 270°C,  at 5°C/min
7 mL/min
1
          111 JCV- I I Ull VUIUIIIG.       1  JU,I-
 DCPA monoacid and diacid metabolites included in method scope; DCPA diacid
 metabolite used for validation studies.  DCPA is a dimethyl ester.
                                 8151  -  25
                                 Revision  0
                             September  1994

-------
                                    TABLE 3
    RETENTION TIMES (MINUTES) OF PFB DERIVATIVES OF CHLORINATED  HERBICIDES
Herbicide
                           Gas Chromatographic Column
Thin-film DB-5a
SP-2250*
Thick-film DB-5C
Dalapon
MCPP
Dicamba
MCPA
Dichloroprop
2,4-D
Silvex
2,4,5-T
Dinoseb
2,4-DB
10.41
18.22
18.73
18.88
19.10
19.84
21.00
22.03
22.11
23.85
12.94
22.30
23.57
23.95
24.10
26.33
27.90
31.45
28.93
35.61
13.54
22.98
23.94
24.18
24.70
26.20
29.02
31.36
31.57
35.97
a  DB-5 capillary column, 0.25 /zm film  thickness, 0.25 mm  ID  x  30  m  long.
   Column temperature, programmed: 70°C for 1 minute, program 10°C/min.  to
   240°C,  hold for 17 minutes.

b  SP-2550 capillary column, 0.25 pirn film  thickness, 0.25  mm  ID x  30 m  long.
   Column temperature, programmed: 70°C for i minute, program 10°C/min.  to
   240°C,  hold for 10 minutes.

c  DB-5 capillary column, 1.0 /urn film thickness, 0.32 mm  ID x 30 m long.
   Column temperature, programmed: 70°C for 1 minute, program 10°C/min.  to
   240°C,  hold for 10 minutes.
                                   8151  -  26
                                                    Revision 0
                                                September 1994

-------
                               TABLE  4
               ACCURACY AND  PRECISION FOR  METHOD  8151
   DIAZOMETHANE DERIVATIZATION, ORGANIC-FREE REAGENT WATER MATRIX
Analyte
Acifluorfen
Bentazon
Chloramben
2,4-D
Dalapon
2,4-DB
DCPA diacidb
Dicamba
3,5-Dichlorobenzoic acid
Dichloroprop
Dinoseb
5-Hydroxydicamba
4-Nitrophenol
Pentachlorophenol
Picloram
2,4,5-TP
2,4,5-T
Spike
Concentration
(M9/L)
0.2
1
0.4
1
10
4
0.2
0.4
0.6
2
0.4
0.2
1
0.04
0.6
0.4
0.2
Mean8 Standard
Percent Deviation of
Recovery Percent Recovery
121
120
111
131
100
87
74
135
102
107
42
103
131
130
91
117
134
15.7
16.8
14.4
27.5
20.0
13.1
9.7
32.4
16.3
20.3
14.3
16.5
23.6
31.2
15.5
16.4
30.8
Mean percent recovery calculated from 7-8 determinations of spiked
organic-free reagent water.

DCPA monoacid and diacid metabolites included in method scope; DCPA
diacid metabolite used for validation studies.  DCPA is a dimethyl ester.
                              8151  -  27
    Revision 0
September 1994

-------
                                  TABLE 5
                  ACCURACY  AND  PRECISION FOR METHOD 8151
                 DIAZOMETHANE DERIVATIZATION, CLAY MATRIX
Analyte
     Mean
Percent Recovery8
   Linear
Concentration
   Rangeb
   (ng/g)
      Percent
     Relative
Standard Deviation0
      (n-20)
Dicamba
MCPP
MCPA
Dichloroprop
2,4-D
2,4,5-TP
2,4,5-T
2,4-DB
Dinoseb
95.7
98.3
96.9
97.3
84.3
94.5
83.1
90.7
93.7
0.52
620
620
1.5
1.2
0.42
0.42
4.0
0.82
- 104
- 61,800
- 61,200
- 3,000
- 2,440
- 828
- 828
- 8,060
- 1,620
7.5
3.4
5.3
5.0
5.3
5.7
7.3
7.6
8.7
   Mean percent recovery calculated from 10 determinations of spiked clay
   and clay/still bottom samples over the linear concentration range.

   Linear concentration range was determined using standard solutions and
   corrected to 50 g solid samples.

   Percent relative standard deviation was calculated using standard
   solutions, 10 samples high in the linear concentration range, and 10
   samples low in the range.
                                8151  - 28
                                               Revision 0
                                           September 1994

-------
                                            TABLE 6
                     RELATIVE RECOVERIES OF PFB DERIVATIVES OF HERBICIDES8
           Standard
         Concentration
Relative recoveries, %
Analyte
MCPP
Dicamba
MCPA
Dichloroprop
2,4-D
Silvex
2,4,5-T
2,4-DB
Mean
mg/L
5.
3.
10.
6.
9.
10.
12.
20.

1
9
1
0
8
4
8
1

1
95.6
91.4
89.6
88.4
55.6
95.3
78.6
99.8
86.8
2
88.8
99.2
79.7
80.3
90.3
85.8
65.6
96.3
85.7
3
97.1
100
87.0
89.5
100
91.5
69.2
100
91.8
4
100
92.7
100
100
65.9
100
100
88.4
93.4
5
95.5
84.0
89.5
85.2
58.3
91.3
81.6
97.1
85.3
6
97.2
93.0
84.9
87.9
61.6
95.0
90.1
92.4
89.0
7
98.1
91.1
92.3
84.5
60.8
91.1
84.3
91.6
87.1
8
98.2
90.1
98.6
90.5
67.6
96.0
98.5
91.6
91.4
Mean
96.3
92.7
90.2
88.3
70.0
93.3
83.5
95.0

Percent recovery determinations made using eight spiked water samples.
                                           8151 - 29
                                               Revision  0
                                           September  1994

-------
                                            METHOD 8151
CHLORINATED HERBICIDES  BY  GC  USING  METHYLATION  OR PENTAFLUOROBENZYLATION
                      DERIVATIZATION:  CAPILLARY  COLUMN  TECHNIQUE

                  Extraction/Hydrolysis  of Waste  and  Soil  Samples
               NO
1

Concentrate and/or
dilute based on
whether derivatization
is by dlazomethane
orPFB
                                                    7.0 Does
                                                  sample con
                                                   tain a hign
                                                    cone of
                                                    waste?
                                                                 72.1.1  Weigh sample
                                                                 and add to beaker;
                                                                 add add and spike;
                                                                 mix wed.
    7.1.1.2
 Does analysis
include herbicide
    esters?
                                                                       72.1.2 Optimize
                                                                       ultrasonic solid extrac
                                                                       Son for each matrix
                   7.2.1.8.1  AddKOHand
                   water  Reflux for 2 hrs.
                   Allow to cod.
                                                                       72.1.3 AddMeCI/
                                                                       acetone to sample 4
                                                                       extract 3 mm ; let
                                                                       serte & decant extract
                  7.2.1.8.2 Transfer the
                  hydrofyzed solutton to a
                  sep funnel and extract 3
                  times with MeCI.
                  Discard extracts.
                                                                       7.2.1.445 Ultra-
                                                                       soracalty extract sample
                                                                       2 more times with MeCI
                  7.2.1.8.3 Acidify and
                  extract 3 times with
                  diethyl efter.  Combine
                  and dry the extracts 2 hrs
                                                                      72.15 Combine organic
                                                                      extracts, centrifuge, and
                                                                      filter extract. Dry for
                                                                      2 hrs
                       Is additional
                        cleanup
                        required?
                                                                      7.2.1.6 Concentrate
                                                                      extract to about 5 ml
                                                                      with Snyder column.
7.2.1.8.4 Proceed to
Section 7.4 to concentrate
extract
               72.1.9.1 Extract 3 times
               with KOH. Discard the
               MeCI.
                                          72.1.9.2 Acidify and
                                          extract 3 times with
                                          diethyl ether. Combine
                                          and dry the extracts 2 hrs
                                                               72.17
                                                            Does analysis
                                                           include herbicide
                                                               esters?
              If hydrolysis is not
              required, proceed to Section
              7.4.4, Nitrogen Slowdown.
                                              8151  -  30
                                                                                        Revision  0
                                                                                  September  1994

-------
                                           METHOD  8151
                                           (continued)

Extraction/Hydrolysis  of  Aqueous  Samples  and  Extract  Concentration
             73.1.1  Measure 1 Lot
             sample and transfer to
             a 2L sep. funnel.
             7.3.1.2 Add2SOgNad
             to sample and shake
             to dissolve
                    7.3.1.3
                 Does analysts
             include hetxoda esters?
           7.3.1.4 Add 12N sulfunc
           acid and shake. Add
           until pH < 2
           7.3.1.5 Adddfettyl
           ether to sample and
           extract Save both
           phases
            73.1.3.1 AddSNNaOHto
            sample and shake. Add
            until pH > 12.  Let stand
            1 hr.
             7.3.1.3.2 AddMeCland
             extract by shaking for
             Zmin.  Discard MeCI.
       73.1.6 Return aqueous phase
       to separately runnel and repeat
       extraction 2 more times, combine
       extracts, and allow extract to
       remain in contact with sodium
       sulfats for 2 hrs.
                    Does
                    difficult
                 emulsion form
                  >1/3 solvent
                   volume?
Employ mechanical techniques
to complete phase separation
(e.g. stirring, filtration through
glass wool, centritugatlon, or
other physical methods).
Discard MeCI.
               7.3.1.3.3  Repeat
               extraction twice more.
               Discard MeCI.
                                         Employ mechanical techniques
                                         to oompMta phase separation
                                         (e.g. stirring, filtration through
                                         glass wool, centritugation. or
                                         other physical methods).
                                         Save both phases.
                                                                                      7.3.1.7 Pour extract
                                                                                      through glass wool and
                                                                                      proceed to Section 7.4.1
                                                                                     7.4.1 Place K-D apparatus
                                                                                     in water bath, concentrate
                                                                                     and cool
7 4.2 • 7.4.4  Complete
concentration with micro-
Snyder column or nitrogen
blow down.
                                           74.5 Dilute extract
                                           with 1 mL isooctane and
                                           0 SmLmethanol
                                            8151  -  31
                                                      Revision  0
                                                September  1994

-------
                                          METHOD  8151
                                           (continued)

                                 Extract  Derlvatization
    7.4.5 Dilute extract
    to 4 ml with acetone
    7.5.2.1  Add potassium
    carbonate and PFBBr.
    Close tube, mix 4 heat
    7.5.2.2 Evaporate with
    nitrogen to 0.5 ml. Add
    2 mL hexane and repeat
     75.23 Redissorve the
     residue in 2 mL toluene:
     hexane (1  6)
    7.5.14  Load sodium
    sulfate / silica cleanup
    column with residue.
          .4.5 Will
       PFB or dlaio-
     metiane deriva
        zatkxibe
          used?
7.4.5 OiluM extract
to 4 mL with dietfiyl
ether
                                          7.5.1
                                         WHIlrw \ DiazaW
                                      Bubbler or the\  Kit
                                       DiazaMKlt
                                       metxidbe
                                         used?
                                        7.5.1  1 Assemble the
                                        diazomethane bubbler
                                             (Rgure 1)
  7.5.1.1.1 Add 5 mL to 1st test
  tube. Add 1 mL dtethyl ether.
  1 mL carbitol, 1.5 mL of 37% KOH
  and 0.1  • 0.2 g DlazakJ to the
  2nd tube.  Bubble wrth nitrogen
  for 10 min or until yellow persists
                                                                                    7.5.1 2  Read kit
                                                                                    instructions
                                                                                 7.5.1.2.1  Add2mL
                                                                                 diazomethane soiuoon
                                                                                 Let stand for 10 min
                                                                                 and swirl
      7.5.1.1.2 Remove con
      centra tor tube and seal
      it Store at room temp.
    7.5.2.5 Bute column
    with enough toluene:
    hexane to collect 8 mL
    eluant
7526 Discard 1st fraction
and continue edition with
enough toluene : hexane (1 :9)
to collect 8 mL more eluant
Transfer to a 10 mL volumetric
flask and dilute to the mark
with hexane
7.5.1.1.3  Add silicic acid to
concentrator tube and let stand
until nitrogen evolution has
stopped. Adjust sample volume
to 10 mL with hexane. Stopper.
Immediate analysis is recommended
       7.5.1.2.2 Rinse ampule with
       diethyl ether and evaporate
       to 2 mL to remove diazomettiane
       Alternatively, silicic acid
       may be added.
       75.1.1.5 If necessary
       store at 4 C in the dark
       for a max of 28 days.
                                                                                7.5.1.2.3 Dilute sample
                                                                                to 10 mL with hexane
      7.6.1 47.6.2 SetGC
      conditions
                                            8151   -   32
                                                                 Revision  0
                                                           September  1994

-------
                            METHOD  8151
                            (continued)

             Analysis  by  Gas  Chromatography
   7.7 Internal or external
   calibration may be used
   (See method 8000).
                                 7.8.1  Add 10 uL internal
                                 standard to the sample
                                 prior to injection.
7.8.2 See method 8000 for
analysis sequence, appropriate
dilutions, establishing daily
retention time windows, and
identification criteria. Check
stds every 10 samples.
   7.84 Record volume
   Injected and the resulting
   peak sizes.
7.8.5 Determine the
identity and quantify
component peaks.
      7.8.6
    Have stds
   and samples
 been prepared and
   analyzed the
        way?
Calculate the correction
for molecular weight of
methyl ester vs herbicide
                                        7.8.6  Calculate con-
                                        centration using procedure
                                        in Method 8000.
7.8.7 Perform further
cleanup if necessary
i

                             8151   -  33
                                     Revision  0
                               September  1994

-------
00
N>
4X
O

-------
                                 METHOD 8240A

       VOLATILE  ORGANICS  BY  GAS  CHROMATOGRAPHY/HASS  SPECTROMETRY  fGC/HS)


1.0   SCOPE AND APPLICATION

      1.1   Method 8240  is  used to determine volatile organic compounds  in a
variety of solid waste matrices.  This  method  is applicable to nearly all  types
of samples,  regardless of water content, including ground water, aqueous sludges,
caustic  liquors,  acid liquors,  waste  solvents,  oily  wastes, mousses,  tars,
fibrous  wastes,  polymeric   emulsions,   filter  cakes,  spent  carbons,  spent
catalysts, soils, and sediments.  The following compounds can be  determined by
this method:
Analyte
CAS No.
  Appropriate Technique
                  Direct
Purge-and-Trap    Injection
Acetone
Acetonitrile
Acrolein
Acrylonitrile
Ally! alcohol
Ally! chloride
Benzene
Benzyl chloride
Bromoacetone
Bromochloromethane (I.S.)
Bromodichloromethane
4-Bromofluorobenzene (surr.)
Bromoform
Bromomethane
2-Butanone
Carbon disulfide
Carbon tetrachloride
Chlorobenzene
Chlorobenzene-de (I.S.)
Chi orodi bromomethane
Chloroethane
2-Chloroethanol
2-ChToroethyl vinyl ether
Chloroform
Chloromethane
Chloroprene
3 -Chi oropropi oni tri 1 e
l,2-Dibromo-3-chloropropane
1,2-Dibromoethane
Oi bromomethane
l,4-Dichloro-2-butene
Dichl orodi fluoromethane
1,1-Dichloroethane
67-64-1
75-05-8
107-02-8
107-13-1
107-18-6
107-05-1
71-43-2
100-44-7
598-31-2
74-97-5
75-27-4
460-00-4
75-25-2
74-83-9
78-93-3
75-15-0
56-23-5
108-90-7
108-90-7
124-48-1
75-00-3
107-07-3
110-75-8
67-66-3
74-87-3
126-99-8
542-76-7
96-12-8
106-93-4
74-95-3
764-41-0
75-71-8
75-34-3
PP
PP
PP
PP
PP
a
a
PP
PP
a
a
a
a
a
PP
PP
a
a
a
a
a
PP
a
a
a
a
ND
PP
a
a
PP
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
pc
pc
a
a
a
a
a
a
                                   8240A -  1
                               Revision 1
                                July 1992

-------
Appropriate Technique

Analyte
1,2-Dichloroethane
1, 2 -Dichl oroethane -djsurr.)
1, 1-Dichloroethene
trans -l,2-Dichloroethenec
1,2-Dichloropropane
l,3-Dichloro-2-propanol
cis-l,3-Dichloropropene
trans-l,3-Dichloropropene
1,2,3,4-Diepoxybutane
1,4-Difluorobenzene (I.S.)
1,4-Dioxane
Epichlorohydrin
Ethanol
Ethyl benzene
Ethylene oxide
Ethyl methacrylate
2-Hexanone
2-Hydroxypropionitrile
lodomethane
Isobutyl alcohol
Malononitrile
Methacrylonitrile
Methyl ene chloride
Methyl iodide
Methyl methacrylate
4-Methyl -2-pentanone
Pentachlproethane
2-Picoline
Propargyl alcohol
B-Propiolactone
Propionitrile
n-Propylamine
Pyridine
Styrene
1,1,1 , 2-Tetrachl oroethane
1,1,2 , 2-Tetrachl oroethane
Tetrachloroethene
Toluene
Toluene-d8 (surr.)
1,1,1-Trichl oroethane
1 , 1 ,2-Trichl oroethane
Trichloroethene
Trichlorofluoromethane
1 , 2 , 3-Tri chl oropropane
Vinyl acetate
Vinyl chloride
Xylene (Total)



CAS No.b
107-06-2
107-06-2
75-35-4
156-60-5
78-87-5
96-23-1
10061-01-5
10061-02-6
1464-53-5
540-36-3
123-91-1
106-89-8
64-17-5
100-41-4
75-21-8
97-63-2
591-78-6
78-97-7
74-88-4
78-83-1
109-77-3
126-98-7
75-09-2
74-88-4
80-62-6
108-10-1
76-01-7
109-06-8
107-19-7
57-57-8
107-12-0
107-10-8
110-86-1
100-42-5
630-20-6
79-34-5
127-18-4
108-88-3
108-88-3
71-55-6
79-00-5
79-01-6
75-69-4
96-18-4
108-05-4
75-01-4
1330-20-7
8240A - 2


Purge-and-Trap
a
a
a
a
a
PP
a
a
a
a
PP
i
i
a
PP
a
PP
NO
a
PP
PP
PP
a
a
a
PP
i
PP
PP
PP
PP
a
i
a
a
a
a
a
a
a
a
a
a
a
a
a
a


Direct
Injection
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
pc
a
a
a
a
a
a
a
a
pc
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
Revision 1
July 1992

-------
    Adequate response by this technique.
    Chemical Abstract Services Registry Number.
    cis- and trans-1,2-dichloroethene  may coelute on packed columns.   For better
    separation of these two compounds use Method 8260.
pp  Poor purging efficiency resulting in high EQLs.
i   Inappropriate technique for this analyte.
pc  Poor chromatographic behavior.

     1.2   Method 8240 can  be used to quantitate most volatile organic compounds
that have boiling points below 200°C and that are insoluble or slightly soluble
in water.  Volatile water-soluble compounds can be included in this analytical
technique.   However,  for  the more soluble  compounds,  quantitation  limits are
approximately ten times higher because of poor purging efficiency.  The method
is also limited to compounds that elute as sharp peaks from a GC column packed
with graphitized carbon lightly  coated with  a carbowax.  Such compounds include
low molecular  weight  halogenated hydrocarbons, aromatics,  ketones,  nitriles,
acetates, acrylates,  ethers,  and  sulfides.  See Table 1 for a list of compounds,
retention times, and  their characteristic  ions that  have  been  evaluated  on a
purge-and-trap GC/MS system.

     1.3   The  estimated   quantitation  limit  (EQL)  of  Method  8240  for  an
individual compound  is  approximately 5 jug/kg  (wet weight)  for soil/sediment
samples, 0,5 mg/kg (wet weight)  for  wastes, and 5 yug/L  for  ground  water  (see
Table 2).  EQLs will  be proportionately higher for sample extracts and samples
that require dilution or reduced sample size  to avoid saturation of  the detector,

     1.4   Method 8240 is based upon a purge-and-trap, gas chromatographic/mass
spectrometric (GC/MS) procedure.   This method is restricted to use by, or under
the supervision of, analysts experienced  in the use of purge-and-trap systems and
gas chromatograph/mass spectrometers,  and  skilled in the interpretation of mass
spectra and their use as a quantitative tool.

     1.5   To increase purging efficiencies of acrylonitrile and  acrolein, refer
to Methods 5030 and 8030 for proper purge-and-trap conditions.


2.0  SUMMARY OF METHOD

     2.1   The volatile compounds are introduced into the gas chromatograph by
the purge-and-trap method or by direct injection (in limited applications).  The
components are  separated via  the gas chromatograph and detected  using a  mass
spectrometer,  which   is  used  to provide  both  qualitative  and  quantitative
information.    The  chromatographic  conditions,   as  well  as  typical  mass
spectrometer operating parameters, are given.

     2.2   If the  above  sample  introduction techniques are  not applicable, a
portion of the sample is dispersed in  methanol to dissolve the volatile organic
constituents.  A portion of the methanolic solution is combined with organic-free
reagent water in a specially designed purging chamber.  It is then analyzed by
purge-and-trap GC/MS following the normal  water method.

     2.3   The  purge-and-trap  process - An  inert  gas is  bubbled through the
solution  at  ambient  temperature,  and the volatile  components  are efficiently
transferred  from  the aqueous phase  to  the vapor  phase.   The  vapor is swept

                                   8240A -  3                        Revision 1
                                                                     July 1992

-------
through a  sorbent  column where  the volatile  components  are trapped.   After
purging is  completed, the sorbent column is heated and backflushed with  inert gas
to  desorb   the  components   onto  a gas  chromatographic  column.    The  gas
chromatographic column  is heated to elute the components, which are detected with
a mass spectrometer.


3.0  INTERFERENCES

     3.1   Interferences  purged  or  coextracted from  the  samples  will  vary
considerably from  source  to source, depending  upon the particular  sample  or
extract being  tested.   The  analytical  system,  however,  should  be  checked  to
ensure freedom from interferences, under the analysis conditions, by analyzing
method blanks.

     3.2   Samples  can be   contaminated  by  diffusion of   volatile  organics
(particularly methylene chloride  and fluorocarbons)  through  the septum  seal into
the sample  during shipment and storage.  A trip blank,  prepared from organic-free
reagent water and carried  through the sampling and handling protocol, can serve
as a check  on such contamination.

     3.3   Cross contamination can  occur whenever  high-concentration  and low-
concentration  samples   are   analyzed   sequentially.    Whenever  an  unusually
concentrated sample  is analyzed,  it   should  be followed  by the analysis  of
organic-free reagent water to check for cross contamination.   The purge-and-trap
system may  require extensive bake-out  and  cleaning  after  a high-concentration
sample.

     3.4   The  laboratory  where volatile  analysis  is  performed  should  be
completely  free of solvents.

     3.5   Impurities   in the purge  gas and  from organic  compounds out-gassing
from the plumbing  ahead of  the  trap account for the majority of contamination
problems.     The   analytical  system  must  be  demonstrated  to   be  free  from
contamination under the conditions  of  the  analysis  by  running reagent blanks.
The use of  non-TFE  plastic coating, non-TFE thread sealants, or flow controllers
with rubber components in the purging device should be avoided.


4.0  APPARATUS AND MATERIALS

     4.1   Microsyringes - 10 /iL, 25 /nl_, 100 /uL, 250 /iL,  500 juL, and  1,000 ^l.
These syringes should  be equipped with a 20 gauge (0.006 in.  ID) needle having
a length sufficient to extend from the  sample inlet  to  within 1  cm of  the glass
frit in the purging device.   The needle length will  depend upon the dimensions
of the purging device  employed.

     4.2   Syringe valve - Two-way, with Luer ends (three each), if applicable
to the purging device.

     4.3   Syringe - 5 ml, gas-tight with shutoff valve.

     4.4   Balances -  Analytical, 0.0001 g, and top-loading,  0.1 g.


                                   8240A -  4                        Revision  1
                                                                      July 1992

-------
     4.5   Glass scintillation vials -  20  ml,  with  screw caps and Teflon liners
or glass culture tubes with a screw cap and Teflon liner.

     4.6   Volumetric flasks,  Class  A  -  10 mL  and 100 ml,  with ground-glass
stoppers.

     4.7   Vials - 2 ml,  for GC autosampler.

     4.8   Spatula - Stainless steel.

     4.9   Disposable pipets - Pasteur.

     4.10  Heater or  heated  oil  bath  - Should  be capable of  maintaining  the
purging chamber to within 1°C over  the  temperature  range of ambient  to  100°C.

     4.11  Purge-and-trap device -  The  purge-and-trap  device  consists of three
separate pieces of  equipment:  the  sample  purger, the trap,  and  the desorber.
Several complete devices  are commercially  available.

           4.11.1 The recommended  purging  chamber is  designed to accept  5 ml
     samples with a  water  column at least 3  cm deep.  The  gaseous headspace
     between the water column and the trap must have a  total volume of less than
     15 ml.  The purge gas  must pass  through the water column as finely divided
     bubbles with a diameter of  less than 3  mm at the origin.   The purge  gas
     must be introduced  no more  than 5 mm from the base  of  the  water column.
     The sample purger,   illustrated  in  Figure 1,  meets  these design criteria.
     Alternate  sample purge devices  may be  utilized,  provided  equivalent
     performance is demonstrated.

           4.11.2 The trap  must  be  at least  25  cm   long  and  have  an  inside
     diameter of  at  least  0.105 in.   Starting  from  the  inlet,  the trap  must
     contain the following  amounts  of adsorbents:  1/3 of 2,6-diphenylene oxide
     polymer, 1/3 of silica gel, and 1/3 of coconut  charcoal.  It is recommended
     that 1.0 cm of methyl  silicone coated packing be  inserted at the inlet to
     extend the life  of  the  trap  (see  Figure 2).   If it is not  necessary to
     analyze  for  dichlorodifluoromethane  or  other  fluorocarbons  of  similar
     volatility, the  charcoal  can  be eliminated and  the  polymer  increased to
     fill  2/3  of the trap.   If only  compounds  boiling  above 35°C  are  to be
     analyzed,  both  the  silica gel  and  charcoal can be  eliminated   and  the
     polymer increased to fill the entire  trap.   Before  initial  use, the trap
     should be conditioned  overnight at 180°C  by backflushing with an inert gas
     flow of at least  20  mL/min.  Vent the  trap effluent to the  room,  not to the
     analytical column.   Prior to daily use, the  trap  should be conditioned for
     10 minutes  at  180°C with backflushing.    The trap may  be vented  to  the
     analytical column during daily conditioning.  However,  the column must be
     run through the temperature program prior to analysis of samples.

           4.11.3 The desorber should be capable  of rapidly heating  the trap to
     180°C  for  desorption.  The polymer  section of the  trap should not be heated
     higher than 180 C, and  the remaining sections should not exceed 220°C during
     bake out mode.   The desorber  design   illustrated in  Figure  2 meets these
     criteria.
                                   8240A  -  5                         Revision 1
                                                                     July 1992

-------
      4.11.4 The purge-and-trap device may  be  assembled as a separate unit
or may be coupled to a gas chromatograph, as shown in Figures 3 and 4.

      4.11.5 Trap Packing Materials

             4.11.5.1    2,6-Diphenylene  oxide  polymer  -  60/80  mesh,
      chromatographic grade (Tenax GC or equivalent).

             4.11.5.2    Methyl   silicone   packing   -    OV-1   (3%)   on
      Chromosorb-W,  60/80 mesh or equivalent.

             4.11.5.3    Silica gel  - 35/60  mesh,  Davison,  grade  15 or
      equivalent.

             4.11.5.4    Coconut charcoal - Prepare from Barnebey  Cheney,
      CA-580-26,  lot  #M-2649,  by crushing through  26  mesh screen  (or
      equivalent).

4.12  Gas chromatograph/mass spectrometer system

      4.12.1 Gas  chromatograph  -  An  analytical  system  complete  with  a
temperature programmable  gas  chromatograph and all  required accessories
including syringes,  analytical columns, and gases.

      4.12.2 Column  - 6 ft x  0.1  in.  ID  glass,  packed with 1% SP-1000 orr
Carbopack-B (60/80 mesh) or equivalent.

      4.12.3 Mass spectrometer - Capable  of  scanning  from 35-260 amu every
3 seconds or less, using 70 volts (nominal)  electron energy  in the electron
impact mode and  producing  a mass  spectrum  that  meets all  the criteria in
Table 3 when 50 ng of 4-bromofluorobenzene (BFB) are injected through the
gas chromatograph inlet.

      4.12.4 GC/MS interface - Any GC-to-MS interface  that gives acceptable
calibration points at 50 ng or less per injection  for each of the analytes
and achieves all  acceptable performance criteria (see  Table  3) may be used.
GC-to-MS  interfaces  constructed   entirely  of  glass or  of  glass-lined
materials  are  recommended.  Glass  can  be deactivated by  silanizing with
di chlorodimethyl si 1ane.

      4.12.5 Data system  -  A  computer system that allows  the continuous
acquisition and  storage  on machine  readable media  of  all  mass  spectra
obtained throughout  the  duration  of the chromatographic  program  must be
interfaced to the mass spectrometer.  The computer must have software that
allows  searching  any GC/MS data  file for  ions  of a specified mass and
plotting such  ion abundances versus time  or  scan number. This type of plot
is defined as an Extracted Ion Current Profile (EICP). Software must also
be available  that  allows integrating the abundances  in  any  EICP  between
specified  time  or  scan  number limits.  The most  recent  version  of the
EPA/NIST Mass Spectral Library should also be available.
                              8240A -  6                        Revision 1
                                                                July 1992

-------
5.0  REAGENTS

     5.1   Reagent grade chemicals  shall be used in all  tests.  Unless otherwise
indicated, it is intended that  all  reagents shall  conform to the specifications
of the Committee on Analytical  Reagents  of the American Chemical Society, where
such specifications are  available.   Other  grades may  be used,  provided it is
first ascertained that the reagent  is  of sufficiently high purity to permit its
use without lessening the accuracy of the determination.

     5.2   Organic-free reagent water -  All  references  to water in this method
refer to organic-free reagent water, as  defined in Chapter One.

     5.3   Stock solutions -  Stock  solutions may be prepared from pure standard
materials or purchased as certified solutions.   Prepare  stock standard solutions
in methanol, using assayed liquids or gases,  as appropriate.

           5.3.1  Place about 9.8 ml of  methanol in a 10 ml tared ground-glass-
     stoppered volumetric  flask.   Allow the  flask to  stand,  unstoppered,  for
     about 10 minutes or until all  alcohol wetted surfaces have dried.   Weigh
     the flask to the nearest 0.0001 g.

           5.3.2  Add the assayed reference  material,  as  described below.

                  5.3.2.1     Liquids -  Using a 100 /iL  syringe,  immediately add
           two or more drops of  assayed reference material to the flask; then
           reweigh.   The liquid must fall directly  into the  alcohol  without
           contacting the neck of the flask.

                  5.3.2.2     Gases - To  prepare  standards for any compounds
           that boil below  30°C (e.g.  bromomethane, chloroethane, chloromethane,
           or vinyl chloride),  fill  a 5 ml  valved gas-tight  syringe  with  the
           reference  standard  to  the  5.0  ml mark.   Lower  the  needle  to 5 mm
           above the methanol meniscus.  Slowly  introduce the reference standard
           above the surface  of the liquid.  The heavy gas will rapidly dissolve
           in the methanol.  Standards may  also be prepared by using a lecture
           bottle  equipped with a  Hamilton  Lecture  Bottle   Septum (#86600).
           Attach Teflon tubing to the side-arm relief valve and  direct a gentle
           stream of gas into the methanol  meniscus.

           5.3.3  Reweigh, dilute to volume, stopper, and then mix by inverting
     the flask  several  times.   Calculate the concentration  in  milligrams  per
     liter (mg/L) from the net  gain  in weight.  When compound purity is assayed
     to  be 96%  or greater,  the weight  may  be  used without  correction  to
     calculate the concentration of the  stock standard.  Commercially prepared
     stock standards may be used at any  concentration if  they are certified by
     the manufacturer or by an independent  source.

           5.3.4  Transfer the stock standard solution  into a  Teflon sealed
     screw cap  bottle.   Store, with  minimal  headspace,  at  -10°C  to -20°C  and
     protect from 1ight.

           5.3.5  Prepare fresh standards every two months for gases.  Reactive
     compounds  such  as  2-chloroethylvinyl  ether  and  styrene  may  need  to be
     prepared more frequently.   All other standards must  be replaced after six

                                   8240A - 7                         Revision 1
                                                                     July 1992

-------
     months.  Both  gas   and  liquid  standards  must  be  monitored  closely by
     comparison to  the  initial calibration curve and by comparison  to QC check
     standards. It may be necessary to replace the  standards  more  frequently if
     either check exceeds a 25% difference.

     5.4   Secondary dilution  standards - Using stock standard solutions, prepare
in methanol, secondary dilution standards containing  the compounds of interest,
either singly or  mixed  together.   Secondary  dilution standards must be stored
with minimal headspace and should be checked frequently for signs of degradation
or evaporation, especially just  prior to preparing calibration standards  from
them.

     5.5   Surrogate  standards  - The  surrogates recommended are toluene-da,
4-bromofluorobenzene, and l,2-dichloroethane-d4.  Other compounds may be used as
surrogates, depending upon the analysis requirements.  A stock surrogate solution
in methanol  should  be prepared as  described in Section  5.3,  and  a surrogate
standard spiking  solution should be prepared from the stock at a concentration
of 250 M9/10 mL  in  methanol.   Each sample  undergoing GC/MS  analysis  must be
spiked with 10 (j.1 of the  surrogate  spiking solution prior  to analys.is.

     5.6   Internal    standards  -   The   recommended   internal   standards  are
bromochloromethane,  1,4-difluorobenzene,  and  chlorobenzene-d5.  Other compounds
may be used as internal standards as long as they have retention times similar
to the compounds  being detected by GC/MS.  Prepare internal standard stock and
secondary  dilution  standards   in  methanol using  the procedures  described in-
Sections 5.3 and  5.4.   It is  recommended that the secondary dilution standard
should be  prepared  at  a  concentration of 25  mg/L of each  internal  standard
compound.  Addition  of 10  y.1 of this standard to 5.0 ml of  sample or calibration
standard would be the equivalent of 50 M9/L.

     5.7   4-Bromofluorobenzene (BFB)  standard - A  standard solution containing
25 ng/jil_ of BFB in  methanol  should  be prepared.

     5.8   Calibration standards -  Calibration  standards  at  a minimum of  five
concentrations should be prepared from the secondary dilution  of stock standards
(see Sections 5.3 and 5.4).   Prepare these  solutions  in  organic-free reagent
water.  One of the concentrations should be at a concentration near,  but above,
the method detection limit. The remaining  concentrations should correspond to the
expected range of  concentrations found in real samples but  should not exceed the
working range of the GC/MS system.   Each standard should contain each  analyte for
detection  by  this  method (e.g. some  or all  of  the  target analytes may be
included). Calibration standards must be prepared daily.

     5.9   Matrix spiking standards  - Matrix spiking standards should  be prepared
from volatile organic compounds which  will be representative of the compounds
being  investigated.      The   suggested   compounds   are   1,1-dichloroethene,
trichloroethene,  chlorobenzene, toluene,  and benzene.  The standard should be
prepared  in  methanol,   with  each  compound  present  at   a  concentration  of
250 /jg/10.0 ml.

     5.10  Great  care must be  taken to maintain the  integrity of all  standard
solutions.  It is  recommended  that  all  standards in methanol  be  stored at -10°C'
to -20°C  in screw  cap amber  bottles with  Teflon liners.


                                   8240A  - 8                         Revision  1
                                                                     July  1992

-------
     5.11  Methanol,  CH3OH.   Pesticide  quality  or equivalent.  Store apart from
other solvents.

     5.12  Reagent Tetraglyme - Reagent tetraglyme is defined as tetraglyme in
which interference is  not observed at the method detection limit  of compounds of
interest.

           5.12.1 Tetraglyme (tetraethylene glycol  dimethyl ether, Aldrich #17,
     240-5 or equivalent),  CgH^Oj. Purify  by treatment at reduced pressure in a
     rotary evaporator. The  tetraglyme  should  have a  peroxide  content of less
     than 5 ppm as indicated by  EM Quant Test Strips (available from Scientific
     Products Co., Catalog No. P1126-8 or equivalent).

           CAUTION:     Glycol ethers  are suspected  carcinogens. All  solvent
                        handling should be  done in a hood while  using proper
                        protective equipment to minimize  exposure to liquid and
                        vapor.

           Peroxides  may be removed by passing the tetraglyme through a column
     of  activated  alumina.  The tetraglyme is  placed  in  a round  bottom flask
     equipped with a  standard taper  joint,  and  the  flask  is affixed to a rotary
     evaporator. The  flask is immersed  in  a water bath at 90-100°C and a vacuum
     is  maintained at < 10  mm Hg  for  at least two  hours  using a  two stage
     mechanical pump.  The  vacuum system  is equipped  with an all  glass trap,
     which is  maintained  in a dry  ice/methanol  bath. Cool  the tetraglyme to
     ambient temperature and add 100 mg/L of 2,6-di-tert-butyl-4-methy1-phenol
     to  prevent peroxide  formation.  Store the tetraglyme in a  tightly sealed
     screw cap bottle in an area that is not contaminated by solvent vapors.

           5.12.2 In  order  to demonstrate  that all interfering  volatiles have
     been removed from the  tetraglyme,  an organic-free reagent water/tetraglyme
     blank must be analyzed.

     5.13  Polyethylene glycol,  H(OCH2CH2)nOH.   Free  of   interferences  at  the
detection limit of the analytes.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1   See  the  introductory material  to  this  chapter,  Organic  Analytes,
Section 4.1.
7.0  PROCEDURE

     7.1   Direct injection -  In very limited applications  (e.g. aqueous process
wastes), direct  injection  of the  sample  into the GC/MS  system  with a  10  /xL
syringe may be  appropriate.   One such application is  for verification  of the
alcohol content  of an  aqueous  sample prior  to  determining if the  sample  is
ignitable (Methods  1010  or 1020).   In this case,  it  is  suggested that  direct
injection be used.  The detection limit is very  high  (approximately 10,000 M9/L);
therefore, it is only permitted when concentrations in  excess of 10,000 M9/L are
expected or for water soluble compounds that do not purge.  The system must be
calibrated by direct injection (bypassing the purge-and-trap device).

                                   8240A -  9                        Revision 1
                                                                     July 1992

-------
7.2   Initial calibration for purge-and-trap procedure

      7.2.1  Recommended GC/MS operating conditions

      Electron energy:               70 volts (nominal).
      Mass range:                    35-260 amu.
      Scan time:                     To  give 5  scans/peak,  but  not  to
                                     exceed 7 sec/scan.
      Initial column temperature:    45°C.
      Initial column holding time:   3 minutes.
      Column temperature program:    8°C/minute.
      Final column temperature:      220°C.
      Final column holding time:      15 minutes.
      Injector temperature:           200-225°C.
      Source temperature:            According     to    manufacturer's
                                     specifications.
      Transfer line temperature:      250-300°C.
      Carrier gas:                   Hydrogen at 50 cm/sec or helium at 30
                                     cm/sec.

      7.2.2  Each GC/MS system must  be  hardware  tuned to meet the criteria
in Table 3 for a 50 ng injection  or  purging of 4-bromofluorobenzene  (2 Mi-
injection  of the  BFB  standard).   Analyses must  not  begin  until  these
criteria are met.

      7.2.3  Assemble a purge-and-trap  device that meets the specification
in Section 4.11.  Condition the trap overnight at 180°C  in the purge mode
with an inert gas flow of at  least 20 mL/min.  Prior to use, condition the
trap daily for 10 min  while backflushing at  180°C with  the column at 220°C.

      7.2.4  Connect the purge-and-trap device to a gas  chromatograph.

      7.2.5  Prepare   the   final   solutions  containing   the   required
concentrations  of  calibration standards,  including  surrogate standards,
directly in the purging device (use freshly prepared stock solutions when
preparing  the  calibration  standards for  the initial  calibration.)   Add
5.0 ml of organic-free reagent water to the purging device.  The organic-
free  reagent  water is  added to   the  purging device  using  a 5 ml  glass
syringe  fitted  with a  15 cm,  20 gauge  needle.   The needle  is  inserted
through the sample inlet shown in Figure 1.  The internal  diameter of the
14 gauge needle that forms the sample inlet will  permit insertion of the 20
gauge needle.  Next, using a  10  yuL  or  25  nl microsyringe  equipped with a
long needle (Section 4.1), take a volume of the secondary dilution solution
containing appropriate concentrations of the calibration standards (Section
5.8).  Add  the aliquot of calibration solution directly to the organic-free
reagent water  in  the  purging device by  inserting  the needle through the
sample inlet.  When discharging the contents of the microsyringe,  be sure
that  the  end of the  syringe needle is  well  beneath  the surface of the
organic-free reagent water.   Similarly,  add  10 ^L of the internal standard
solution (Section 5.6).  Close the 2 way syringe  valve  at the sample inlet.

      7.2.6  Carry out the purge-and-trap analysis procedure as described
in Section 7.4.1.
                             8240A - 10                        Revision 1
                                                                July 1992

-------
      7.2.7  Tabulate the  area  response of the  characteristic  ions  (see
Table  1)   against  concentration  for  each  compound  and  each  internal
standard.   Calculate response factors  (RF)  for  each compound relative to
one of  the internal  standards.    The  internal standard  selected  for the
calculation of the RF for a compound should be the internal standard that
has  a  retention  time  closest  to the  compound  being  measured  (Section
7.5.2).   The RF is calculated as follows:
where:

      Ax     =     Area of the characteristic  ion  for the compound being
                   measured.
      Ajs     =     Area of the characteristic ion for the specific internal
                   standard.
      Cis     =     Concentration of the specific internal  standard.
      Cx     =     Concentration of the compound being measured.

      7.2.8  The average RF must be calculated for each compound.  A system
performance check  should  be  made before this  calibration  curve is used.
Five  compounds  (the System  Performance Check  Compounds,  or  SPCCs)  are
checked  for a  minimum  average  response  factor.    These compounds  are
chloromethane,  1,1-dichloroethane,  bromoform,  1,1,2,2-tetrachloroethane,
and chlorobenzene.  The minimum acceptable average RF for these compounds
should be 0.300 (0.250  for bromoform) .   These compounds typically have RFs
of 0.4-0.6  and  are used  to  check  compound instability and  to check for
degradation caused  by  contaminated  lines or active  sites  in the system.
Examples' of these occurrences are:

             7.2.8.1     Chloromethane  - This compound is the most likely
      compound to be lost  if the purge  flow is too fast.

             7.2.8.2     Bromoform - This compound is one of  the compounds
      most likely to be purged very poorly if the purge flow is too slow.
      Cold  spots  and/or  active  sites  in the  transfer lines may adversely
      affect  response.    Response  of the  quantitation  ion  (m/z  173) is
      directly  affected  by  the  tuning  of  BFB  at  ions   m/z  174/176.
      Increasing the m/z  174/176 ratio  may improve bromoform response.

             7.2.8.3     Tetrachloroethane and 1,1-dichloroethane  - These
      compounds are degraded by contaminated transfer lines  in purge-and-
      trap  systems and/or  active sites  in trapping materials.

      7.2.9  Using  the RFs  from  the initial  calibration,  calculate the
percent relative standard deviation (%RSD)  for  Calibration Check Compounds
(CCCs).

              SD
      %RSD  = - x 100
where:
                             8240A  -  11                         Revision  1
                                                                July  1992

-------
      RSO    =     relative standard deviation.
      x      =     mean of 5 initial RFs for a compound.
      SO     =     standard deviation of average RFs for a compound.
N (x,-
£
Ni=l N
-x)2
- 1
      SO


      The %RSD  for  each individual CCC  should  be less than  30  percent.
This criterion must be met in order for  the  individual  calibration  to be
valid.  The CCCs are:

      1,1-Dichloroethene,
      Chloroform,
      1,2-Dichloropropane,
      Toluene,
      Ethyl benzene,  and
      Vinyl chloride.

7.3   Daily GC/MS calibration

      7.3.1  Prior to  the analysis  of samples, inject or purge  50 ng of the
4-bromofluorobenzene standard.  The resultant mass spectra  for  the BFB must
meet all of the  criteria given in  Table  3  before sample analysis begins,.
These criteria must be demonstrated each 12 hour shift.

      7.3.2  The initial calibration curve  (Section 7.2) for each compound
of interest must be checked  and  verified once  every  12 hours  of analysis
time.  This is accomplished by analyzing  a calibration  standard that is at
a concentration  near  the midpoint  concentration  for  the working  range of
the GC/MS by checking the SPCC (Section 7.3.3)  and CCC  (Section 7.3.4).

      7.3.3  System  Performance  Check  Compounds  (SPCCs)  -  A  system
performance check must  be made  each  12 hours.    If the SPCC  criteria are
met, a comparison of response factors  is made for all compounds.   This is
the  same  check  that is applied  during  the initial  calibration.   If the
minimum response factors are not met,  the system must  be evaluated,  and
corrective action must be taken before sample analysis begins.  The minimum
response factor  for volatile SPCCs is  0.300  (0.250  for Bromoform).   Some
possible problems are  standard mixture  degradation,  injection port  inlet
contamination, contamination  at the front end of the analytical column, and
active sites in the column or chromatographic system.

      7.3.4  Calibration  Check  Compounds  (CCCs):     After  the  system
performance check is  met, CCCs listed  in Section 7.2.9 are used  to  check
the validity of the initial calibration.  Calculate the  percent difference
using:

                      RF,  - RFC
      % Difference =  	=	  x  100
                         RF,
                             8240A - 12                        Revision 1
                                                                July 1992

-------
where:

      RF     =     average  response   factor   from  initial  calibration
                   (Section 7.2).
      RF     =     response   factor   from  current   verification  check
                   standard.

      If the percent difference  for  any  compound is  greater than 20,  the
laboratory should consider this a warning limit.  If the percent difference
for each CCC  is  less  than 25%,  the  initial calibration  is  assumed  to be
valid.  If the criterion  is not  met  (> 25% difference),  for any one CCC,
corrective action MUST be  taken.   Problems similar to  those listed  under
SPCCs could affect  this  criterion.   If  no source of the problem can be
determined  after corrective  action   has  been  taken,  a  new five  point
calibration  MUST  be  generated.    This  criterion  MUST  be  met  before
quantitative sample analysis begins.

      7.3.5  The  internal  standard responses  and  retention  times in  the
check calibration standard must  be evaluated  immediately  after  or during
data acquisition.  If  the  retention time  for any  internal standard changes
by more than 30 seconds  from the  last  daily calibration (Section 7.3),  the
chromatographic system must be inspected for malfunctions and corrections
must  be  made,  as required.   If  the  EICP area  for  any of  the  internal
standards changes by a factor  of  two  (-  50% to +  100%) from the last daily
calibration standard check, the  mass spectrometer must be  inspected  for
malfunctions  and  corrections  must  be  made,  as  appropriate.    When
corrections are made,  reanalysis of samples analyzed while the system was
malfunctioning are necessary.

7.4   GC/MS analysis

      7.4.1  Water samples

             7.4.1.1     Screening of  the  sample prior  to purge-and-trap
      analysis  will   provide  guidance  on whether  sample  dilution  is
      necessary  and  will  prevent  contamination  of the  purge-and-trap
      system.    Two  screening techniques  that  can  be  used  are:    the
      headspace  sampler   (Method  3810)  using  a  gas chromatograph   (GC)
      equipped with a  photo ionization detector  (PID)  in  series with an
      electrolytic conductivity  detector  (HECD);  and  extraction of  the
      sample with hexadecane and.analysis of the extract  on a GC with a FID
      and/or an ECD (Method 3820).

             7.4.1.2     All samples  and standard solutions must be allowed
      to warm to ambient temperature before analysis.

             7.4.1.3     Set  up  the  GC/MS system  as outlined  in Section
      7.2.1.

             7.4.1.4     BFB  tuning criteria  and daily  GC/MS calibration
      criteria must be met (Section 7.3) before analyzing samples.

             7.4.1.5     Adjust the  purge  gas (helium)  flow rate to  25-
      40 ml/min on the purge-and-trap  device.   Optimize  the flow rate to

                             8240A - 13                        Revision  1
                                                                July  1992

-------
provide the best response for chloromethane and bromoform, if these
compounds are analytes.   Excessive  flow rate  reduces chloromethane
response, whereas insufficient flow reduces bromoform response (see
Section 7.2.8).

       7.4.1.6     Remove the plunger from a 5 ml syringe  and attach
a closed syringe valve.   Open  the  sample or  standard bottle, which
has been allowed to come to ambient temperature, and carefully pour
the sample  into the syringe  barrel  to just short  of overflowing.
Replace  the  syringe plunger  and compress  the  sample.    Open  the
syringe valve and vent  any  residual  air while  adjusting the sample
volume to  5.0 ml.   This process of taking an  aliquot destroys  the
validity of  the liquid sample  for  future analysis;  therefore,  if
there is only one VOA vial,  the analyst  should fill  a second syringe
at this time to  protect against  possible loss  of sample integrity.
This  second  sample is  maintained only until   such  time  when  the
analyst  has  determined  that  the  first  sample  has  been  analyzed
properly.  Filling one  20  mL syringe would allow the use of only one
syringe.  If a second  analysis is needed from a syringe, it must be
analyzed within  24  hours.   Care must be taken  to  prevent air from
leaking into the syringe.

       7.4.1.7     The  following  procedure   is   appropriate  for
diluting purgeable  samples.  All  steps must be performed without
delays until the diluted sample is in a gas tight syringe.

             7.4.1.7.1    Dilutions may be made in volumetric flasks
       (10 to 100 mL).   Select the volumetric flask that will allow
       for the  necessary  dilution.    Intermediate  dilutions  may be
       necessary for extremely large dilutions.

             7.4.1.7.2   Calculate the approximate volume of organic-
       free  reagent water  to  be  added  to  the   volumetric  flask
       selected and add slightly less  than this quantity of organic-
       free reagent water to the flask.

             7.4.1.7.3   Inject the proper aliquot of samples from
       the  syringe  prepared  in Section 7.4.1.6  into  the  flask.
       Aliquots of  less than  1 mL are not recommended.   Dilute the
       sample to the mark with organic-free reagent water.  Cap the
       flask, invert,  and  shake  three  times.  Repeat above procedure
       for additional  dilutions.

             7.4.1.7.4   Fill  a 5 ml  syringe with the diluted sample
       as in Section 7.4.1.6.

       7.4.1.8     Add 10.0 p.1 of surrogate spiking  solution (Section
5.5)  and 10 /*L  of internal  standard  spiking  solution (Section 5.6)
through the  valve  bore of the syringe;  then close the  valve.   The
surrogate and internal  standards may be mixed and added as a single
spiking  solution.   The addition of 10  pi  of  the surrogate spiking
solution to  5  mL  of   sample  is equivalent  to a  concentration of
50 jjg/L of each  surrogate standard.
                       8240A - 14                        Revision 1
                                                          July 1992

-------
       7.4.1.9     Attach the syringe-syringe valve assembly to the
syringe valve on  the  purging device.  Open  the  syringe  valves and
inject the sample into the purging chamber.

       7.4.1.10    Close both valves and  purge  the  sample for 11.0 ±
0.1 minutes at ambient temperature.

       7.4.1.11     At the conclusion  of  the  purge  time,  attach the
trap to the chromatograph,  adjust  the  device  to the desorb mode, and
begin  the  gas chromatographic  temperature  program and  GC/MS  data
acquisition. Concurrently,  introduce the trapped materials to the gas
chromatographic column  by  rapidly heating the trap  to 180°C  while
backflushing  the  trap with  inert  gas between 20 and  60  mL/min for
4 minutes.  If this  rapid heating  requirement cannot be met, the gas
chromatographic column must  be used as a secondary trap  by cooling it
to  30°C   (or  subambient,   if   problems  persist)  instead  of  the
recommended initial  program temperature of 45°C.

       7.4.1.12    While the  trap is  being  desorbed  into  the gas
chromatograph, empty the purging  chamber.  Wash  the  chamber with a
minimum of two  5 mL  flushes  of organic-free  reagent  water  (or
methanol  followed by organic-free  reagent water)  to avoid carryover
of pollutant compounds into subsequent analyses.

       7.4.1.13    After  desorbing   the   sample  for  4  minutes,
recondition the trap by  returning the purge-and-trap  device to the
purge mode.   Wait 15  seconds;  then close the  syringe  valve on the
purging device  to  begin gas   flow  through  the trap.   The  trap
temperature should be maintained  at 180°C.   Trap temperatures up to
220 C may  be employed;  however,  the higher temperature will shorten
the useful  life  of the trap.  After approximately  7 minutes, turn off
the  trap  heater and open  the  syringe valve to  stop  the  gas  flow
through the trap.  When  cool, the  trap is ready for the next sample.

       7.4.1.14    If the initial  analysis  of a sample  or a dilution
of  the sample  has  a concentration  of  analytes that exceeds the
initial calibration range,  the  sample  must be reanalyzed  at a higher
dilution.   Secondary ion quantitation  is  allowed  only when there are
sample interferences with the primary  ion.  When a sample  is analyzed
that  has  saturated  ions from  a  compound,  this analysis  must be
followed by  a blank organic-free  reagent  water analysis.   If the
blank  analysis  is not  free of interferences,  the system  must be
decontaminated.   Sample  analysis may not  resume until a blank can be
analyzed that is free of interferences.

       7.4.1.15    For matrix spike analysis, add 10 juL of the matrix
spike  solution  (Section 5.9) to  the  5  ml of  sample  to  be purged.
Disregarding any dilutions,  this is equivalent  to a concentration of
50 jig/L of each matrix  spike standard.

       7.4.1.16    All  dilutions  should keep  the  response  of the
major constituents (previously saturated  peaks)  in  the  upper half of
the  linear range of the  curve.  Proceed to Sections 7.5.1 and 7.5.2
for  qualitative and quantitative  analysis.

                       8240A -  15                        Revision  1
                                                          July  1992

-------
      7.4.2  Water miscible liquids

             7.4.2.1     Water  miscible  liquids  are  analyzed  as  water
      samples after first diluting them at least 50 fold with organic-free
      reagent water.

             7.4.2.2     Initial  and  serial  dilutions  can  be prepared by
      pipetting  2  mL  of the  sample  to  a  100 mi  volumetric  flask  and
      diluting  to   volume  with  organic-free  reagent  water.    Transfer
      immediately to a 5 mL gas tight syringe.

             7.4.2.3     Alternatively, prepare dilutions directly in a 5
      ml syringe filled with  organic-free  reagent water by adding at least
      20 /iL,  but not more than  100 /uL of liquid sample.   The sample is
      ready for addition of internal  and surrogate standards.

      7.4.3  Sediment/soil and  waste  samples - It is  highly recommended
that all samples of this  type be screened prior to the purge-and-trap GC/MS
analysis.  The headspace method (Method  3810)  or the hexadecane extraction
and screening method  (Method  3820)  may be used for this  purpose.   These
samples may  contain percent quantities  of purgeable organics  that  will
contaminate the  purge-and-trap  system,  and  require extensive cleanup and
instrument downtime.   Use the  screening data  to  determine  whether to use
the low-concentration  method  (0.005-1  mg/kg) or the  high-concentration
method  (> 1 mg/kg).

             7.4.3.1     Low-concentration method - This  is designed for
      samples containing individual purgeable compounds of < 1 mg/kg.  It
      is limited to  sediment/soil  samples  and waste  that  is of a similar
      consistency  (granular and porous).  The low-concentration method is
      based on  purging  a  heated  sediment/soil  sample mixed with organic-
      free reagent water containing the surrogate and internal standards.
      Analyze all reagent blanks and standards under the same conditions as
      the  samples.    See Figure  5  for an  illustration of a  low  soils
      impinger.

                   7.4.3.1.1   Use  a   5   g   sample   if   the  expected
             concentration is  <  0.1  mg/kg or  a  1  g sample  for expected
             concentrations between 0.1 and 1 mg/kg.

                   7.4.3.1.2   The GC/MS  system  should  be  set  up  as in
             Sections 7.4.1.2-7.4.1.4.  This  should  be done prior to the
             preparation  of  the sample to  avoid  loss  of  volatiles  from
             standards and samples.- A heated  purge calibration curve must
             be  prepared and  used  for the  quantitation  of  all  samples
             analyzed  with   the   low-concentration method.     Follow  the
             initial  and  daily calibration instructions, except for the
             addition of a 40°C purge  temperature.

                   7.4.3.1.3   Remove the  plunger from a 5 mL Luerlock type
             syringe  equipped  with  a  syringe  valve  and  fill  until
             overflowing with water.   Replace  the plunger and compress the
             water to vent trapped air.  Adjust the volume to  5.0 mL.  Add
             10  juL each of surrogate spiking  solution  (Section 5.5) and

                             8240A - 16                        Revision 1
                                                                July 1992

-------
internal   standard  solution  (Section  5.6)   to  the  syringe
through the valve.   (Surrogate spiking solution and internal
standard  solution may be  mixed  together.)  The addition of 10
Ml of the surrogate spiking solution to 5 g of sediment/soil
is equivalent to 50 jug/kg of each surrogate standard.

      7.4.3.1.4   The sample  (for volatile organics) consists
of the  entire  contents  of  the  sample container.   Do  not
discard any  supernatant  liquids.    Mix the contents of  the
sample container with a  narrow  metal  spatula.    Weigh  the
amount determined  in Section  7.4.3.1.1  into a tared  purge
device.  Note and record  the  actual weight to the nearest  0.1
9-

      7.4.3.1.5   Determine  the  percent dry  weight of  the
soil/sediment sample.  This  includes  waste  samples that  are
amenable  to percent dry  weight determination.   Other wastes
should be reported  on a wet-weight basis.

            7.5.3.1.5.1 Immediately after weighing  the sample
      for extraction,  weigh 5-10 g of  the sample into a tared
      crucible.  Determine the % dry weight of the sample by
      drying  overnight  at  105°C.     Allow  to  cool  in  a
      desiccator  before  re-weighing.    Concentrations   of
      individual analytes  are  reported relative to  the  dry
      weight of sample.

            WARNING:     The drying  oven  should  be  contained
                        in a  hood  or vented.   Significant
                        laboratory contamination may result
                        from a heavily contaminated hazardous
                        waste sample.

            % dry weight = g of dry sample x 100
                              g of sample

      7.4.3.1.6   Add the  spiked water to  the purge device,
which contains the weighed amount of sample, and connect the
device to the purge-and-trap system.

      NOTE: Prior to the attachment of the purge device,  the
            procedures  in  Sections 7.4.3.1.4  and  7.4.3.1.6
            must   be    performed    rapidly   and    without
            interruption to avoid loss of volatile organics.
            These  steps  must  be  performed  in  a   laboratory
            free of solvent fumes.

      7.4.3.1.7   Heat the sample to 40°C ± 1°C and purge the
sample for 11.0 +.0.1 minute.

      7.4.3.1.8   Proceed  with  the analysis  as  outlined in
Sections 7.4.1.11-7.4.1.16.  Use 5 ml of the  same organic-free
reagent water  as in  the reagent blank.   If saturated peaks
occurred or  would  occur if a  1 g  sample  were analyzed,  the

                8240A -  17                         Revision  1
                                                   July  1992

-------
       high-concentration method must be followed.

             7.4.3.1.9   For  low-concentration  sediment/soils  add
       10 nl of the matrix spike solution (Section 5.9) to the 5 mL
       of  organic-free   reage'nt  water  (Section  7.4.3.1.3).    The
       concentration for a 5 g sample would  be equivalent to 50 jug/kg
       of each matrix spike standard.

       7.4.3.2     High-concentration method - The method  is based on
extracting the sediment/soil with methanol.  A waste sample is either
extracted  or  diluted,  depending on  its  solubility   in  methanol.
Wastes  (i.e.  petroleum  and  coke  wastes)  that  are   insoluble  in
methanol are diluted with reagent tetraglyme or possibly polyethylene
glycol  (PEG).    An  aliquot  of the extract is added  to organic-free
reagent  water  containing internal  standards.   This  is   purged  at
ambient temperature.   All samples with an expected concentration of
> 1.0 mg/kg should be analyzed by this method.

             7.4.3.2.1   The sample  (for volatile organics) consists
       of  the  entire contents  of the  sample   container.   Do  not
       discard  any  supernatant  liquids.   Mix  the contents of  the
       sample    container  with   a   narrow  metal   spatula.     For
       sediment/soil  and solid wastes that are insoluble in methanol,
       weigh 4 g (wet weight) of sample into a tared 20 ml vial.  Use
       a top loading balance.   Note  and record  the actual  weight to
       0.1 gram and determine the percent dry weight  of  the sample
       using the procedure  in Section 7.4.3.1.5.  For waste that is
       soluble  in  methanol,  tetraglyme,  or PEG, weigh  1 g  (wet
       weight)  into a tared scintillation vial  or culture tube or a
       10 ml volumetric flask.  (If a vial  or tube is used, it must
       be calibrated prior to use.  Pipet 10.0 mL of solvent into the
       vial and mark the  bottom of the meniscus.    Discard  this
       solvent.)

             7.4.3.2.2   For sediment/soil  or solid  waste,  quickly
       add 9.0  mL  of appropriate solvent, then add  1.0  mL  of  the
       surrogate  spiking solution  to  the  vial.    For   a  solvent
       miscible  sample, dilute  the  sample  to  10  mL  with  the
       appropriate  solvent  after adding  1.0 mL  of  the  surrogate
       spiking  solution.  Cap and shake for 2 minutes.

             NOTE: Sections   7.4.3.2.1   and   7.4.3.2.2   must   be
                   performed  rapidly and  without interruption  to
                   avoid loss-  of volatile organics.   These  steps
                   must  be  performed  in  a  laboratory   free  from
                   solvent  fumes.

             7.4.3.2.3   Pipet approximately  1  mL of the extract to
       a  GC  vial  for  storage,   using  a  disposable  pipet.    The
       remainder may be disposed of.   Transfer approximately 1 mL of
       appropriate  solvent  to a  separate  GC vial  for use  as  the
       method blank for each set of samples.   These extracts may be
       stored at 4°C  in  the dark, prior to analysis.  The addition of
       a  100   /LtL  aliquot  of  each   of  these  extracts  in  Section

                       8240A - 18                        Revision  1
                                                          July 1992

-------
             7.4.3.2.6 will  give a  concentration equivalent to 6,200
             of each surrogate standard.

                   7.4.3.2.4   The  GC/MS  system should  be  set up  as  in
             Sections 7.4.1.2-7.4.1.4.  This  should  be  done  prior to the
             addition of the solvent extract to organic-free reagent water.

                   7.4.3.2.5   Table 4 can  be  used to determine the volume
             of solvent extract to  add to the  5 ml of organic-free reagent
             water for  analysis.   If a screening procedure  was  followed
             (Method 3810  or  3820),  use  the  estimated  concentration  to
             determine  the  appropriate  volume.   Otherwise,  estimate the
             concentration range of the sample from the low-concentration
             analysis to determine the appropriate volume.   If the sample
             was submitted as  a high-concentration sample,  start  with 100
             /iL.   All  dilutions  must  keep  the  response of the  major
             constituents (previously saturated peaks) in the  upper half of
             the linear range  of the curve.

                   7.4.3.2.6   Remove the plunger from  a 5.0 mL  Luerlock
             type syringe  equipped with a  syringe  valve and  fill  until
             overflowing with  water.   Replace  the plunger and compress the
             water to vent trapped  air.  Adjust the volume to  4.9 ml.  Pull
             the plunger back  to 5.0 mL to allow volume for the addition of
             the sample extract and of  standards.  Add  10 /iL of  internal
             standard solution.  Also add  the volume of solvent  extract
             determined in  Section  7.4.3.2.5 and a volume of extraction or
             dissolution solvent to  total  100 nl  (excluding methanol  in
             standards).

                   7.4.3.2.7   Attach the syringe-syringe valve assembly to
             the syringe valve  on  the purging device.   Open the  syringe
             valve  and  inject  the  organic-free  reagent  water/methanol
             sample into the purging chamber.

                   7.4.3.2.8   Proceed with the  analysis as  outlined  in
             Section 7.4.1.11-7.4.1.16.  Analyze all reagent  blanks on the
             same instrument as that  use for  the  samples.  The standards
             and blanks should also contain 100 nl of solvent to  simulate
             the sample conditions.

                   7.4.3.2.9   For a matrix spike in the high-concentration
             sediment/soil   samples,  add 8.0  ml of  methanol,  1.0  ml  of
             surrogate spike solution (Section 5.5),  and 1.0  ml of matrix
             spike solution (Section  5.9)  as  in Section 7.4.3.2.2.   This
             results in a 6,200 jig/kg concentration  of  each  matrix spike
             standard when  added to a 4  g sample.  Add a 100 /xL aliquot of
             this extract to 5 ml of organic-free reagent water for purging
             (as per Section 7.4.3.2.6).
7.5   Data interpretation

      7.5.1  Qualitative analysis
                             8240A - 19                        Revision 1
                                                                July 1992

-------
       7.5.1.1     An  analyte  (e.g.  those  listed  in  Table  1)  is
identified by comparison of  the  sample  mass  spectrum with the mass
spectrum of a standard  of the suspected  compound (standard reference
spectrum).  Mass spectra for  standard reference should be obtained on
the user's GC/MS within  the  same 12  hours as  the sample analysis.
These standard reference  spectra  may be  obtained through analysis of
the calibration standards.  Two criteria must be satisfied to verify
identification:   (1)  elation of sample component  at the  same  GC
relative retention  time (RRT) as those of the standard component; and
(2) correspondence  of the sample component  and the  standard component
mass spectrum.

             7.5.1.1.1    The  sample  component RRT must compare within
       ±0.06 RRT units of the RRT  of the  standard  component.   For
       reference,  the standard must be run within  the same 12 hours
       as  the sample.    If   coelution  of  interfering  components
       prohibits accurate assignment of the sample  component RRT from
       the total  ion  chromatogram,  the RRT  should  be  assigned  by
       using extracted ion current  profiles for ions unique to the
       component of interest.

             7.5.1.1.2   (1)  All  ions present  in  the standard mass
       spectra  at   a  relative  intensity  greater  than   10%  (most
       abundant ion in the spectrum equals  100% must be present in
       the sample  spectrum).   (2) The relative intensities of ions
       specified in (1) must  agree within plus or minus 20% between
       the standard and sample spectra.   (Example:  For an ion with
       an abundance of 50% in the standard  spectra, the corresponding
       sample abundance must  be between 30 and 70 percent.

       7.5.1.2     For samples containing components not associated
with the calibration standards,  a library  search may be made for the
purpose of tentative identification.  The necessity to perform this
type of  identification will  be  determined by  the type  of analyses
being conducted.  Guidelines  for making tentative identification are:

       (1)   Relative  intensities  of major  ions  in  the  reference
spectrum  (ions > 10% of the most abundant ion) should be present in
the sample spectrum.

       (2)   The relative intensities of the major ions should agree
within + 20%.  (Example:   For an  ion with an abundance of 50% in the
standard  spectrum,  the  corresponding sample ion  abundance  must be
between 30 and 70%).

       (3)   Molecular ions present  in the reference spectrum should
be present in the sample spectrum.

       (4)   Ions  present  in the  sample  spectrum  but  not  in the
reference  spectrum  should  be   reviewed  for  possible  background
contamination or presence of coeluting  compounds.

       (5)   Ions present  in the  reference  spectrum but not in the
sample spectrum should  be  reviewed for possible subtraction from the

                       8240A - 20                        Revision  1
                                                          July 1992

-------
sample  spectrum  because of  background  contamination  or coeluting
peaks.  Data system library reduction programs can sometimes create
these discrepancies.

       Computer generated  library  search  routines should  not use
normalization routines that would misrepresent the library or unknown
spectra when compared  to each other.  Only after visual  comparison of
sample  with  the nearest  library  searches  will  the  mass spectral
interpretation specialist assign a tentative identification.

7.5.2  Quantitative analysis

       7.5.2.1     When  a   compound  has  been   identified,   the
quantitation  of that  compound  will be based  on the   integrated
abundance  from  the   EICP   of  the  primary   characteristic  ion.
Quantitation will  take place using the internal standard  technique.
The internal  standard  used  shall  be the one  nearest  the retention
time of that of a given analyte (e.g. see Table 5).

       7.5.2.2     Calculate the  concentration of  each   identified
analyte in the sample as follows:

       Water and Water-Miscible Waste:


       concentration (ng/l) =
                               (A,.)(RF)(V0)

where:

       Ax    *     Area  of characteristic  ion for  compound being
                   measured.
       Is    =     Amount of internal standard injected (ng).
       Ais    =     Area  of  characteristic  ion   for  the  internal
                   standard.
       RF    =     Response  factor  for  compound   being  measured
                   (Section 7.2.7).
       V0    =»     Volume   of   water  purged  (ml),   taking   into
                   consideration any dilutions made.

       Sediment/Soil, Sludge, and Waste:

       High-concentration:

                                 (A, )(!,)  (Vt)
       concentration (jug/kg)
                                (A,.)(RF)(Vf)(H,)


       Low-concentration:

                                 (Ax )(I.)
       concentration (jig/kg)
                                (Ais)(RF)(Ws)

                       8240A - 21                        Revision  1
                                                          July  1992

-------
                  Ax,  Is, Ajs,  RF =  Same as  in  water  and  water-miscible waste
                                    above.
           where :
                  Vt    =     Volume of total extract (pi) (use 10,000 juL or a
                              factor of this when dilutions are made).
                  Vf    =     Volume of extract added (/iL) for purging.
                  Ws    =     Weight of  sample  extracted or purged  (g).   The
                              wet weight or  dry weight  may be  used,  depending
                              upon the specific applications of the data.

                  7.5.2.3     Sediment/soil samples  are  generally reported on a
           dry weight  basis,  while sludges  and wastes  are reported  on a  wet
           weight basis.   The percent dry weight of the sample (as calculated in
           Section 7.4.3.1.5)  should be reported along with the data in either
           instance.

                  7.5.2.4     Where applicable, an estimate of concentration for
           noncalibrated components in the  sample  should be made.  The formulae
           given above should  be used with  the following  modifications:   The
           areas A  and Ais should be from the total  ion chromatograms, and the
           RF for the  compound  should be assumed to be 1.   The concentration
           obtained should  be  reported indicating   (1) that the  value  is an
           estimate and   (2)  which internal  standard  was  used to  determine
           concentration.    Use  the  nearest   internal   standard   free  of
           interferences.
8.0  QUALITY CONTROL

     8.1   Each  laboratory  that uses these  methods  is required  to  operate a
formal quality control program.  The minimum requirements of this program consist
of an initial  demonstration  of laboratory capability and an ongoing analysis of
spiked samples  to  evaluate  and document quality data.   The  laboratory should
maintain records to document  the quality of  the  data  generated.   Ongoing data
quality checks are compared with established performance criteria to determine
if the results of analyses meet the performance characteristics of the method.
When results of  sample  spikes indicate  atypical  method performance,  a quality
control reference sample should be analyzed to confirm that the measurements were
performed in an  in-control mode of operation.

     8.2   Before processing any samples, the analyst should demonstrate, through
the  analysis  of a calibration  blank,  that  interferences  from  the analytical
system, glassware, and reagents are under control.   Each time a set of samples
is  extracted  or  there  is  a  change in  reagents,  a  reagent blank  should be
processed as a  safeguard  against  chronic  laboratory  contamination.  The blank
samples  should  be carried  through  all   stages  of   sample preparation  and
measurement.

     8.3   The experience of the analyst performing  GC/MS analyses  is invaluable
to the success of the methods.  Each day that analysis is performed, the daily
calibration standard  should be evaluated to determine if the chromatographic
system is operating properly.   Questions that should be asked  are:   Do the peaks
look normal?;  Is the response  obtained  comparable to the  response  from previous

                                  8240A - 22                        Revision  1
                                                                     July  1992

-------
calibrations?   Careful  examination  of the  standard  chromatogram  can indicate
whether the column is still  useable,  the injector is leaking,  the injector septum
needs  replacing,  etc.    If  any changes  are made  to  the system  (e.g.  column
changed), recalibration of the system should take place.

     8.4   Required instrument QC is found in the following section:

           8.4.1  The  GC/MS   system  should  be   tuned  to  meet   the   BFB
     specifications in Section 7.2.2.

           8.4.2  There should be an initial  calibration  of the GC/MS system as
     specified in Section 7.2.

           8.4.3  The GC/MS system  should meet the  SPCC  criteria  specified in
     Section 7.3.3 and the CCC criteria in Section 7.3.4, each 12  hours.

     8.5   To  establish  the   ability  to  generate  acceptable  accuracy  and
precision, the analyst should perform the following operations.

           8.5.1  A  quality  control   (QC)   reference  sample  concentrate  is
     required containing each  analyte at a concentration of 10 mg/L in methanol.
     The QC  reference  sample concentrate may  be  prepared from pure standard
     materials  or purchased  as  certified   solutions.    If  prepared  by  the
     laboratory, the  QC reference  sample  concentrate should be made using stock
     standards prepared independently from those used for calibration.

           8.5.2  Prepare a  QC reference sample  to contain  20  M9/L  of  each
     analyte by adding 200 /A of  QC reference  sample  concentrate  to 100 ml of
     water.

           8.5.3  Four 5 ml aliquots of the well  mixed QC reference sample are
     analyzed according to the method beginning in Section 7.4.1.

           8.5.4  Calculate the average recovery (x)  in M9/U and  the standard
     deviation of the  recovery (s)  in M9/L,  for  each analyte using the  four
     results.

           8.5.5  For  each  analyte  compare s  and x  with the  corresponding
     acceptance criteria_for precision  and  accuracy,  respectively,  found in
     Table 6.   If  s  and x for all  analytes  meet  the  acceptance  criteria, the
     system performance is acceptable and analysis of actual samples can_begirr.
     If any individual s exceeds the precision limit or any individual x falls
     outside the range  for accuracy,  then the system performance is unacceptable
     for that analyte.

           NOTE:  The large number of analytes  in Table 6 present a substantial
                  probability that  one  or more will  fail at  least  one of the
                  acceptance  criteria when  all analytes  of  a  given method are
                  determined,

           8.5.6  When one or more of the analytes tested fail at  least one of
     the acceptance criteria,  the analyst  should  proceed according to Section
     8.5.6.1 or 8.5.6.2.
                                  8240A - 23                        Revision 1
                                                                     July 1992

-------
                  8.5.6.1     Locate and correct the source of the problem and
           repeat the test for all  analytes beginning with Section 8.5.2.

                  8.5.6.2     Beginning with Section 8.5.2, repeat the test only
           for those analytes that  failed to meet criteria.  Repeated failure,
           however,  will  confirm a  general  problem with the measurement system.
           If this  occurs,  locate  and  correct  the  source of  the  problem and
           repeat the test for all  compounds of  interest beginning with Section
           8.5.2.

     8.6   The laboratory should, on an ongoing basis,  analyze a reagent blank
and  a  spiked replicate  for  each  analytical   batch  (up  to  a  maximum of  20
samples/batch) to assess  accuracy.   For  soil and waste samples where detectable
amounts of organics  are present, replicate samples  may be appropriate in place
of spiked replicates.  For laboratories  analyzing one to ten samples per month,
at least one spiked  sample per month is required.

           8.6.1  The  concentration  of  the  spike  in  the  sample  should  be
     determined  as follows:

                  8.6.1.1     If, as in compliance monitoring, the concentration
           of a  specific  analyte  in  the  sample is  being checked  against  a
           regulatory concentration limit,  the spike should be at that limit or
           1 to  5 times higher than the background  concentration determined in
           Step  8.6.2, whichever concentration would be larger.

                  8.6.1.2     If the concentration of  a  specific analyte  in a
           water sample is not being checked against a  specific limit, the spike
           should be  at  20 jig/L or  1 to  5  times  higher  than  the  background
           concentration determined in  Section  8.6.2,  whichever concentration
           would  be   larger.     For  other  matrices,   recommended  spiking
           concentration is 10 times the EQL.

           8.6.2  Analyze one 5-mL  sample  aliquot  to  determine the background
     concentration  (B)   of  each analyte.    If  necessary,  prepare  a  new  QC
     reference sample concentrate  (Step  8.5.1)  appropriate for the background
     concentration in the sample.   Spike a second 5-mL  sample  aliquot with  10 \il
     of the  QC  reference sample concentrate  and analyze  it to  determine the
     concentration after  spiking (A)  of  each  analyte.   Calculate each percent
     recovery (p) as 100(A-B)%/T, where T is the known  true value of the spike.

           8.6.3  Compare the percent recovery  (p)  for each analyte in a water
     sample  with  the corresponding QC  acceptance  criteria found in  Table 6.
     These acceptance criteria were calculated to include an allowance for error
     in measurement  of both the  background  and spike concentrations, assuming a
     spike to background ratio of 5:1.   This error will be  accounted for to the
     extent  that  the analyst's  spike  to background ratio  approaches  5:1.  If
     spiking was  performed  at a concentration lower  than 20  ng/L, the analyst
     should  use  either  the  QC  acceptance criteria  presented  in Table  6,  or
     optional  QC  acceptance  criteria  calculated  for   the  specific   spike
     concentration.   To calculate optional  acceptance criteria  for the recovery
     of  an  analyte:  (1)   Calculate  accuracy  (x') using  the  equation  found in
     Table  7, substituting the  spike  concentration  (T) for C;  (2)  calculate
     overall precision (S') using  the equation  in Table 7, substituting x' for

                                   8240A  - 24                        Revision  1
                                                                     July 1992

-------
     x;  (3)  calculate the  range  for recovery  at  the  spike  concentration as
     (lOOx'/T) ± 2.44(100S'/T)%.

           8.6.4  If  any  individual  p  falls outside  the  designated  range for
     recovery,  that   analyte  has   failed  the  acceptance  criteria.    A  check
     standard  containing  each  analyte  that  failed  the  criteria  should  be
     analyzed as described in Section 8.7.

     8.7   If any analyte in  a  water sample fails  the acceptance criteria for
recovery in  Section  8.6,  a QC  reference  sample containing each  analyte that
failed should be prepared and analyzed.

     NOTE:  The frequency for the required analysis of a QC  reference sample will
           depend upon the number of analytes being simultaneously tested, the
           complexity  of  the  sample  matrix,   and   the  performance  of  the
           laboratory.   If  the entire list  of analytes in Table  6  should be
           measured  in  the  sample  in  Section  8.6,  the probability  that  the
           analysis of a QC  reference sample will be required  is high.  In this
           case the QC reference  sample  should be  routinely analyzed with the
           spiked sample.

           8.7.1  Prepare the  QC  reference sample by adding  10 |iL  of the QC
     reference sample concentrate  (Section  8.5.1 or 8.6.2)  to 5 ml of reagent
     water.   The QC  reference sample needs only to contain the analytes that
     failed criteria  in the test in Section 8.6.

           8.7.2  Analyze the  QC reference sample to determine  the concentration
     measured  (A)  of each analyte.   Calculate each  percent  recovery (ps) as
     100(A/T)%, where T is the true value of the standard concentration.

           8.7.3  Compare the percent  recovery (ps) for each  analyte with the
     corresponding QC acceptance criteria found in  Table 6.  Only analytes that
     failed the test  in Step  8.6  need  to  be compared  with these criteria.  If
     the recovery of  any  such  analyte  falls outside  the designated range, the
     laboratory performance for that  analyte is judged to be  out  of control, and
     the problem should  be immediately identified and corrected.  The  result for
     that analyte in the unspiked  sample  is  suspect and  may  not  be reported for
     regulatory compliance purposes.

     8.8   As part  of the QC program for the laboratory, method  accuracy for each
matrix studied should be assessed and records should be maintained.  After the
analysis of five spiked samples (of the  same matrix) as in Section 8.6, calculate
the average  percent  recovery  (p)  and  the  standard  deviation  of  the percent
recovery (s ).  Express  the accuracy assessment as a percent recovery  interval
from p - 2s   to p  +  2s  .   If  p = 90%  and  s  =  10%,  for example,  the  accuracy
interval  is  expressed as 70-110%.   Update  the accuracy assessment  for each
analyte  on  a  regular  basis   (e.g.  after  each  five  to  ten new   accuracy
measurements).

     8.9   To determine acceptable accuracy and precision limits for surrogate
standards the following procedure should be performed.
                                  8240A -25                        Revision 1
                                                                     July 1992

-------
           8.9.1  For each sample  analyzed,  calculate  the  percent recovery of
     each surrogate in the sample.

           8.9.2  Once a minimum of thirty samples  of the same matrix have been
     analyzed,  calculate the average percent recovery (p) and standard deviation
     of the percent recovery (s) for each of the surrogates.

           8.9.3  For a  given  matrix, calculate  the  upper and  lower  control
     limit for method performance for each surrogate standard.   This should be
     done as follows:

           Upper Control Limit (UCL) = p + 3s
           Lower Control Limit (LCL) = p - 3s

           8.9.4  For aqueous and  soil matrices,  these  laboratory established
     surrogate  control  limits  should,  if  applicable,  be  compared with  the
     control limits listed in Table 8.  The limits given in Table 8 are multi-
     laboratory performance  based  limits  for soil  and  aqueous  samples,  and
     therefore, the single-laboratory limits  established  in Step 8.9.3 should
     fall within those given in Table 8 for these matrices.

           8.9.5  If recovery  is not within limits,  the following procedures are
     required.

           •      Check  to be  sure  there  are  no errors  in  calculations,
                  surrogate  solutions and  internal standards.    Also,  check
                  instrument performance.

           •      Recalculate the data and/or  reanalyze the extract if any of
                  the above checks reveal a problem.

           •      Reextract and reanalyze the sample if none of the above are a
                  problem or flag the data as "estimated concentration".

           8.9.6  At a minimum, each laboratory should update surrogate recovery
     limits on a matrix-by-matrix basis,  annually.

     8.10  It  is  recommended  that  the  laboratory  adopt  additional  quality
assurance practices for use with this method.  The specific practices that are
most productive depend upon the needs of  the laboratory and the nature of the
samples.   Field duplicates  may be  analyzed to assess the precision  of the
environmental measurements. When doubt exists over the  identification of a peak
on the chromatogram, confirmatory techniques such as gas chromatography with a
dissimilar column or a different ionizatioji mode using a mass spectrometer should
be used.  Whenever  possible,  the  laboratory should analyze standard reference
materials and participate  in relevant performance evaluation studies.


9.0  METHOD PERFORMANCE

     9.1   This method was tested  by 15 laboratories using  organic-free reagent
water, drinking water, surface water, and industrial wastewaters  spiked at six
concentrations over the  range 5-600  fj,g/l.   Single operator precision,  overall
precision,  and method  accuracy  were found  to  be directly  related   to the

                                  8240A - 26                        Revision  1
                                                                     July 1992

-------
concentration o* r.he ana'yte and essentially independent of the sample matrix.
Linear equations to describe these relationships are presented in Table 7.


10.0 REFERENCES

1.   U.S. EPA 40 CFR Part  136.  "Guidelines  Establishing Test Procedures for the
     Analysis of Pollutants Under the  Clean Water Act, Method 624," October 26,
     1984.

2.   U.S.  EPA  Contract  Laboratory  Program,  Statement  of Work  for  Organic
     Analysis, July 1985,  Revision.

3.   Bellar, T.A., and J.J. Lichtenberg, J. Amer.  Water  Works  Assoc.,  66(12),
     739-744, 1974.

4.   Bellar, T.A., and J.J. Lichtenberg, "Semi-Automated Headspace Analysis of
     Drinking  Waters  and  Industrial   Waters  for  Purgeable Volatile  Organic
     Compounds," in Van Hall,  ed., Measurement  of  Organic  Pollutants in Water
     and Wastewater, ASTM STP 686, pp. 108-129,  1979.

5.   Budde, W.L. and J.W. Eichelberger, "Performance Tests for the Evaluation of
     Computerized   Gas   Chromatography/Mass   Spectrometry   Equipment   and
     Laboratories,"  EPA-600/4-79-020,  U.S.  Environmental   Protection  Agency,
     Environmental Monitoring  and  Support  Laboratory,  Cincinnati,  Ohio 45268,
     April 1980.

6.   Eichelberger, J.W.,  L.E.  Harris,  and W.L.  Budde,  "Reference  Compound to
     Calibrate Ion Abundance Measurement in Gas Chromatography-Mass Spectrometry
     Systems," Analytical  Chemistry,  47, 995-1000,  1975.
7.   "Method Detection Limit for Methods 624 and 625," Olynyk,  P.,  W.L. Budde,
     and J.W. Eichelberger, Unpublished report,  October 1980.

8.   "Interlaboratory Method Study for EPA  Method 624-Purgeables," Final Report
     for EPA Contract 68-03-3102.

9.   "Method Performance Data for Method 624," Memorandum from  R. Slater and T.
     Pressley,  U.S. Environmental  Protection Agency,  Environmental  Monitoring
     and Support Laboratory, Cincinnati, Ohio 45268, January 17, 1984.

10.  Gebhart,  J.E.;  Lucas,  S.V.; Naber,  S.J.; Berry,  A.M.;   Danison,  T.H.;
     Burkholder, H.M.  "Validation of  SW-846 Methods 8010, 8015, and 8020"; U.S.
     Environmental  Protection  Agency,  Environmental  Monitoring  and  Support
     Laboratory, Cincinnati, Old 45268, July 1987,  Contract No. 68-03-1760.

11.  Lucas,  S.V.;  Kornfeld,  R.A. "GC-MS Suitability  Testing  of RCRA Appendix
     VIII  and Michigan  List  Analytes  ";  U.S.  Environmental Protection Agency,
     Environmental  Monitoring  and Support Laboratory,  Cincinnati,  OH 45268,
     February 20,  1987, Contract No.  68-03-3224.
                                  8240A - 27                        Revision  1
                                                                     July  1992

-------
                                   TABLE  1.
        RETENTION TIMES AND CHARACTERISTIC IONS FOR VOLATILE COMPOUNDS
Compound
Retention
Time (minutes)
Primary Ion  Secondary Ion(s)
Ethylene oxide
Chloromethane
Dichlorodifluoromethane
Bromomethane
Vinyl chloride
Acetonitrile
Chloroethane
Methyl iodide
Methylene chloride
Carbon disulfide
Trichlorofluoromethane
Propionitrile
Ally! chloride
1,1-Dichloroethene
Bromochloromethane (I.S.)
Ally! alcohol
trans- 1,2-Dichloroethene
1,2-Dichloroethane
Propargyl alcohol
Chloroform
l,2-Dichloroethane-d4(surr)
2-Butanone
Methacrylonitrile
Dibrornomethane
2-Chloroethanol
b-Propiolactone
Epichlorohydrin
1,1,1-Tri chloroethane
Carbon tetrachloride
1,4-Dioxane
Isobutyl alcohol
Bromodi Chloromethane
Chloroprene
l,2:3,4-Diepoxybutane
1,2-Dichloropropane
cis-l,3-Dichloropropene
Bromoacetone
Trichloroethene
Benzene
trans-l,3-Dichloropropene
1,1,2-Trichloroethane
3-Chloropropionitrile
1,2-Dibromoethane
Pyridine
1.30
2.30
2.47
3.10
3.80
3.97
4.60
5.37
6.40
7.47
8.30
8.53
8.83
9.00
9.30
9.77
10.00
10.10
10.77
11.40
12.10
12.20
12.37
12.53
12.93
13.00
13.10
13.40
13.70
13.70
13.80
14.30
14.77
14.87
15.70
15.90
16.33
16.50
17.00
17.20
17.20
17.37
18.40
18.57
44
50
85
94
62
41
64
142
84
76
101
54
76
96
128
57
96
62
55
83
65
72
41
93
49
42
57
97
117
88
43
83
53
55
63
75
136
130
78
75
97
54
107
79
44, 43, 42
52, 49
85, 87, 101, 103
96, 79
64, 61
41, 40, 39
66, 49
142, 127, 141
49, 51, 86
76, 78, 44
103, 66
54, 52, 55, 40
76, 41, 39, 78
61, 98
49, 130, 51
57, 58, 39
61, 98
64, 98
55, 39, 38, 53
85, 47
102
43, 72
41, 67, 39, 52, 66
93, 174, 95, 172, 176
49, 44, 43, 51, 80
42, 43, 44
57, 49, 62, 51
99, 117
119, 121
88, 58, 43, 57
43, 41, 42, 74
85, 129
53, 88, 90, 51
55, 57, 56
62, 41
77, 39
43, 136, 138, 93, 95
95, 97, 132
52, 71
77, 39
83, 85, 99
54, 49, 89, 91
107, 109, 93, 188
79, 52, 51, 50
                                  8240A - 28
                                        Revision 1
                                         July 1992

-------
                                   TABLE 1.
                                  (Continued)
Compound
Retention
Time (minutes)
Primary Ion  Secondary Ion(s)
2-Chloroethyl vinyl ether
2-Hydroxypropionitrile
1,4-Difluorobenzene (I.S.)
Malononitrile
Methyl methacrylate
Bromoform
1,1,1 , 2-Tetrachl oroethane
l,3-Oichloro-2-propanol
1,1,2 , 2-Tetrachl oroethane
Tetrachloroethene
1,2,3-Trichloropropane
l,4-Dichloro-2-butene
n-Propylamine
2-Picoline
Toluene
Ethyl methacrylate
Chlorobenzene
Pentachl oroethane8
Ethyl benzene
l,2-Dibromo-3-chloropropane
4-Bromofluorobenzene (surr.)
Benzyl chloride
Styrene
Acetone
Acrolein
Acrylonitrile
Chlorobenzene-dc (I.S.)
Chlorodibromomethane
1,1-Dichloroethane
Ethanol
2-Hexanone
lodomethane
4-Methyl -2-pentanone
Toluene-do (surr.)
Vinyl acetate
Xylene (Total)
18.60
18.97
19.60
19.60
19.77
19.80
20.33
21.83
22.10
22.20
22.20
22.73
23.00
23.20
23.50
23.53
24.60
24.83
26.40
27.23
28.30
29.50
30.83
--

--
::
--
-.

--
--
--
-.
• -
63
44
114
66
69
173
131
79
83
164
75
75
59
93
92
69
112
167
106
157
95
91
104
43
56
53
117
129
63
31
43
142
43
98
43
106
65,106
44,43,42,53
63,88
66,39,65,38
69,41,100,39
171,175,252
131,133,117,119,95
79,43,81,49
85,131,133
129,131,166
75,77,110,112,97
75,53,77,124,89
59,41,39
93,66,92,78
91,65
69,41,99,86,114
114,77
167,130,132,165,169
91
157,75,155,77
174,176
91,126,65,128
104,103,78,51,77
58
55,58
52,51
82,119
208,206
65,83
45,27,46
58,57, 100
127,141
58,57,100
70,100
86
91
a  The base peak at m/e 117 was not  used due to an interference at that mass with
   a  nearly coeluting  internal  standard,  chlorobenzene-d5.
                                  8240A  -  29
                                        Revision 1
                                         July 1992

-------
                         TABLE 2.
ESTIMATED QUANTITATION LIMITS (EQL) FOR VOLATILE ORGANICS3
                           Estimated
                          Quantitation
                            Limits"
Ground water
Volatiles p.g/1
Acetone
Acetonitrile
Allyl chloride
Benzene
Benzyl chloride
Bromodichloromethane
Bromoform
Bromomethane
2-Butanone
Carbon disulfide
Carbon tetrachloride
Chlorobenzene
Chi orodi bromomethane
Chloroethane
2-Chloroethyl vinyl ether
Chloroform
Chloromethane
Chloroprene
l,2-Dibromo-3-chloropropane
1,2-Dibromoethane
Di bromomethane
l,
-------
                                   TABLE 2.
                                  (Continued)
                                      Estimated
                                     Quantitation
                                       Limits6
                            Ground water        Low Soil/Sediment
   Volatiles                    ng/l                  M9/kg
Propionitrile
Styrene
1,1,1,2-Tetrachloroethane
1,1,2,2-Tetrachloroethane
Tetrachloroethene
Toluene
1,1,1-Trichloroethane
1,1,2-Trichloroethane
Trichloroethene
1,2,3-Trichloropropane
Vinyl acetate
Vinyl chloride
Xylene (Total)
100
5
5
5
5
5
5
5
5
5
50
10
5
100
5
5
5
5
5
5
5
5
5
50
10
5
a Sample EQLs are highly matrix dependent.  The EQLs listed herein are provided
  for guidance and may not always  be  achievable.  See the following information
  for further guidance on matrix dependent EQLs.

b EQLs  listed  for soil/sediment  are  based on wet  weight.   Normally  data  is
  reported on a dry weight basis;  therefore,  EQLs will  be higher, based on the
  percent dry weight of each sample.
               Other Matrices                      Factor0
               Water miscible liquid was'te             50
               High-concentration soil and sludge     125
               Non-water miscible waste               500


  CEQL =  [EQL for  low  soil  sediment  (Table  2)]  X  [Factor],  For non-aqueous
          samples, the factor is on a wet weight basis.
                                  8240A - 31                        Revision 1
                                                                     July 1992

-------
                                TABLE 3.
                     BFB KEY ION ABUNDANCE CRITERIA
    Mass             Ion Abundance Criteria
    50               15 to 40% of mass 95
    75               30 to 60% of mass 95
    95               base peak, 100% relative abundance
    96               5 to 9% of mass 95
   173               0% to less than 2% of mass 174
   174               greater than 50% of mass 95
   175               5 to 9% of mass 174
   176               greater than 95% but less than 101% of mass 174
   177               5 to 9% of mass 176
                                TABLE 4.
          QUANTITY OF METHANOL EXTRACT REQUIRED FOR ANALYSIS
                 OF HIGH-CONCENTRATION SOILS/SEDIMENTS
       Approximate                               Volume of
   Concentration Range                        Methanol  Extract*
      500- 10,000 ng/kg                            100  /*L
    1,000- 20,000 Mg/kg                             50  ML
    5,000-100,000 Mg/kg                             10  ML
   25,000-500,000 Mg/kg                            100  ML of 1/50 dilution"
Calculate  appropriate  dilution  factor  for  concentrations  exceeding  this
table.

a  The volume of methanol  added to  5 mL of water being purged should be kept
   constant. Therefore, add to the  5 mL syringe whatever volume of methanol
   is necessary to maintain a volume of 100 ML added to the syringe.

b  Dilute  and  aliquot  of  the methanol  extract  and then take 100 ML  for
   analysis.
                               8240A - 32                        Revision 1
                                                                  July 1992

-------
                                   TABLE 5.
       VOLATILE INTERNAL STANDARDS WITH CORRESPONDING ANALYTES ASSIGNED
                               FOR QUANTITATION
Bromochloromethane

Acetone
Acrolein
Acrylonitrile
Bromomethane
Carbon disulfide
Chloroethane
Chloroform
Chloromethane
Di chlorodi fluoromethane
1,1-Dichloroethane
1,2-Dichloroethane
l,2-Dichloroethane-d4 (surrogate)
1,1-Dichloroethene
trans-1,2-Dichloroethene
lodomethane
Methylene chloride
Tri chlorof1uoromethane
Vinyl chloride
1,4-Difluorobenzene

Benzene
Bromodichloromethane
Bromoform
2-Butanone
Carbon tetrachloride
Chlorodi bromomethane
2-Chloroethyl vinyl ether
Dibromomethane
l,4-Dichloro-2-butene
1,2-Dichloropropane
cis^l,3-Dichloropropene
trans-l,3-Dichloropropene
1,1,1-Trichloroethane
1,1,2-Trichloroethane
Trichloroethene
Vinyl acetate
                        Chlorobenzene-dc
                        Bromofluorobenzene  (surrogate)
                        Chlorobenzene
                        Ethyl benzene
                        Ethyl  methacrylate
                        2-Hexanone
                        4-Methyl-2-pentanone
                        Styrene
                        1,1,2,2-Tetrachloroethane
                        Tetrachloroethene
                        Toluene
                        Toluene-d8  (surrogate)
                        1,2,3-Trichloropropane
                        Xylene
                                  8240A  - 33
                      Revision  1
                       July  1992

-------
                                TABLE 6.
                 CALIBRATION  AND  QC  ACCEPTANCE  CRITERIA3


Parameter
Benzene
Bromodi Chloromethane
Bromoform
Bromomethane
Carbon tetrachloride
Chlorobenzene
2-Chloroethylvinyl ether
Chloroform
Chloromethane
Dibromochloromethane
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
1,1-Dichloroethane
1,2-Dichloroethane
1,1-Dichloroethene
trans -1,2-Dichloroethene
1,2-Dichloropropane
cis-l,3-Dichloropropene
trans-l,3-Dichloropropene
Ethyl benzene
Methylene chloride
1,1,2,2-Tetrachloroethane
Tetrachloroethene
Toluene
1,1,1-Trichloroethane
1,1,2-Trichloroethane
Trichloroethene
Trichlorofluoromethane
Vinyl chloride
Range
for Q
(M9/L)
12.8-27.2
13.1-26.9
14.2-25.8
2.8-37.2
14.6-25.4
13.2-26.8
0-44.8
13.5-26.5
D-40.8
13.5-26.5
12.6-27.4
14.6-a5.4
12.6-27.4
14.5-25.5
13.6-26.4
10.1-29.9
13.9-26.1
6.8-33.2
4.8-35.2
10.0-30.0
11.8-28.2
12.1-27.9
12.1-27.9
14.7-25.3
14.9-25.1
15.0-25.0
14.2-25.8
13.3-26.7
9.6-30.4
0.8-39.2
Q = Concentration measured in QC check
s = Standard deviation of four recovery
x = Average recovery
p, ps = Percent recovery
D = Detected; result
for four recovery
measured.
Limi t
for s
(M9/L)
6.9
6.4
5.4
17.9
5.2
6.3
25.9
6.1
19.8
6.1
7.1
5.5
7.1
5.1
6.0
9.1
5.7
13.8
15.8
10.4
7.5
7.4
7.4
5.0
4.8
4.6
5.5
6.6
10.0
20.0
sample,
Range
for x
(M9/L)
15.2-26.0
10.1-28.0
11.4-31.1
D-41.2
17.2-23.5
16.4-27.4
D-50.4
13.7-24.2
D-45.9
13.8-26.6
11.8-34.7
17.0-28.8
11.8-34.7
14.2-28.4
14.3-27.4
3.7-42.3
13.6-28.4
3.8-36.2
1.0-39.0
7.6-32.4
17.4-26.7
D-41.0
13.5-27.2
17.0-26.6
16.6-26.7
13.7-30.1
14.3-27.1
18.5-27.6
8.9-31.5
D-43.5
in ng/L.
Range
P,PS

37-151
35-155
45-169
0-242
70-140
37-160
0-305
51-138
0-273
53-149
18-190
59-156
18-190
59-155
49-155
0-234
54-156
0-210
0-227
17-183
37-162
D-221
46-157
64-148
47-150
52-162
52-150
71-157
17-181
D-251

measurements, in /ug/L.
measurements, in /ug/L.



must be greater than zero.
Criteria from 40 CFR Part 136 for Method 624 and were calculated assuming a
QC check sample concentration of 20 /ug/L-  These criteria are based directly
upon the method performance data  in Table 7.   Where necessary, the limits for
recovery  have   been  broadened  to  assure applicability  of  the limits  to
concentrations  below those used to develop Table 7.
                               8240A - 34
Revision 1
 July 1992

-------
                                   TABLE 7.
         METHOD ACCURACY AND PRECISION AS FUNCTIONS OF CONCENTRATION3
Parameter
Accuracy, as
recovery, x'
  (M9/L)
Single analyst Overall
precision,  s '  precision.
   (M9/L)      S'  (Mg/L)
Benzene
Bromodi chl oromethane
Bromoform
Bromomethane
Carbon tetrachloride
Chlorobenzene
Chloroethane
2-Chloroethyl vinyl ether3
Chloroform
Chloromethane
Di bromochl oromethane
l,2-Dich1orobenzeneb
1,3-Dichlorobenzene
1,4-Di chlorobenzene6
1,1-Dichloroethane
1,2-Dichloroethane
1,1-Dichloroethene
trans-l,2,-Dichloroethene
l,2-Dich1oropropanea
cis-l,3-Dichloropropenea
trans -1,3-Di chl oropropene3
Ethyl benzene
Methyl ene chloride
1,1,2,2-Tetrachloroethane
Tetrachloroethene
Toluene
1,1,1-Trichloroethane
1,1,2-Trichloroethane
Trichloroethene
Tri chl orof 1 uoromethane
Vinyl chloride
0.93C+2.00
1.03C-1.58
1.18C-2.35
l.OOC
1.10C-1.68
0.98C+2.28
1.18C+0.81
l.OOC
0.93C+0.33
1.03C-1.81
1.01C-0.03
0.94C+4.47
1.06C+1.68
0.94C+4.47
1.05C+0.36
1.02C+0.45
1.12C+0.61
1.05C+0.03
l.OOC
l.OOC
l.OOC
0.98C+2.48
0.87C+1.88
0.93C+1.76
1.06C+0.60
0.98C+2.03
1.06C+0.73
0.95C+1.71
1.04C+2.27
0.99C+0.39
l.OOC
0.26X-1.74
0.15X+0.59
0.12X+0.34
0.43x
0.12X+0.25
0.16X-0.09
0.14X+2.78
0.62x
0.16X+0.22
0.37X+2.14
0.17x-0.18
0.22X-1.45
0.14X-0.48
0.22X-1.45
0.13X-0.05
0.17X-0.32
O.Ux+1.06
0.14X+0.09
0.33x
0.38x
0.25x
0.14X+1.00
O.lBx+1.07
0.16X+0.69
0.13X-0.18
O.lBx-0.71
0.12X-0.15
0.14X+0.02
0.13X+0.36
0.33X-1.48
0.48x
0.25X-1.33
0.20x+1.13
0.17x+1.38
0.58x
O.llx+0.37
0.26x-1.92
0.29X+1.75
0.84x
O.lSx+0.16
0.58x+0,43
0.17X+0.49
O.SOx-1.20
O.lSx-0.82
fl.30x-l.20
0.16X+0.47
0.21X-0.38
0.43X-0.22
0.19X+0.17
0.45x
0.52X
0.34X
0.26X-1.72
0.32X+4.00
0. 20X+0.41
0.16X-0.45
0.22X-1.71
0.21X-0.39
0.18X+0.00
0.12X+0.59
0.34X-0.39
0.65x
x'    =      Expected  recovery  for  one  or  more  measurements  of  a  sample
            containing a concentration of C,  in M9/L.
sr'    =      Expected single analyst  standard  deviation of measurements  at  an
            average concentration of x,  in  /ig/L.
S'    =      Expected interlaboratory standajd deviation of measurements  at  an
            average concentration found of x,  in /zg/L.
C     =      True value for the concentration,  in Mg/L.
x     =      Average recovery  found  for  measurements  of samples  containing  a
            concentration of C,  in ng/L.
a     Estimates  based  upon  the performance  in  a  single  laboratory.
b     Due  to chromatographic  resolution  problems,  performance  statements  for
      these  isomers  are  based  upon the  sums of  their  concentrations.
                                  8240A - 35
                                        Revision 1
                                         July 1992

-------
                                   TABLE 8.
     SURROGATE  SPIKE  RECOVERY  LIMITS FOR WATER AND  SOIL/SEDIMENT SAMPLES
                              Low/High                      Low/High
Surrogate Compound              Water                       Soil/Sediment


4-Bromofluorobenzene           86-115                        74-121
1,2-Dichloroethane-d,           76-114                        70-121
Toluene-dfl                     88-110                        81-117
                                  8240A  - 36                        Revision 1
                                                                     July 1992

-------
                            FIGURE  1.
                        PURGING  CHAMBER
        OPTIONAL
        FOAM TRAP  jH        Exit * lnch °- °

Inch 0. 0. Exit
   10 mm Glass Frit
   Medium PorotitY
                               14 mm 0. 0.
                               Inlet 'A Inch 0. 0.
                                Sample Inlet
                                2-Way Syringe Valve
                                17 cm, 20 Gauge Syringe Needle
                                6 mm 0. D. Rubber Septum
                                  — 10 mm 0 D.
Inlet
4 Inch 0. D.
                                                        1/16 Inch 0 D
                                                        Stainless Ste«:
                                                        13x Molecular
                                                        Sieve Purge
                                                        Gas Filter
                                                        Purge Gas
                                                        Flow Control
                            8240A  -  37
                                Revision  1
                                 July 1992

-------
                           FIGURE  2.
         TRAP PACKINGS  AND  CONSTRUCTION  TO  INCLUDE
             DESORB  CAPABILITY  FOR METHOD  8240A
          Packing Procedure
Construction
Glass Wool   5 mm
               t
  Silica G«l   Bern
   Ttn»x   15cm
3%OV-1    lemj
Glut Wool   5 mm
             Compression
             Pitting Nut
             and Ferrules
             14 Ft. ?n/Foot
             Rtsittanet Wire
             Wrapptd Solid
              Thirmocouple/
              Controiltr
                 Electronic
                 Ttmptratun
                 Control *nd
             Tubing 25 cm
             0.105 In. l.D.
             0.125 In. O.D.
             Stiinlw Stt«l
                 Trap Inlet
                          8240A - 38
                            Revision 1
                              July  1992

-------
                               FIGURE  3.
SCHEMATIC OF  PURGE-AND-TRAP DEVICE - PURGE MODE FOR METHOD  8240A
            CMMMOAS
            *JOM CONT1XX
                                                   4TOMV
                                              TO OCTtCTOA
                                              »

                                             *»- AHUtmCAl OOUIMN
                                             HOTl
                                             AU. UM
                                             ANO OC SHOULD M HCATW
                              8240A - 39
Revision 1
 July  1992

-------
                               FIGURE  4.
SCHEMATIC OF PURGE-ANO-TRAP  DEVICE  -  OESOR8 MODE  FOR METHOD  8240A
                                r- uouo MJCCTOH "ours

                                      CCUAMOCM
                                            fOOCTKTOI

                                             AKH.TTCA1. CtXUMM
                                           AM) (K SHOUU) M HCAHO
                                           TOWX.
                             8240A  -  40
Revision 1
 July  1992

-------
                   FIGURE 5.
              LOW SOILS  IMPINGER
 PU*GE INLET rrrnwo
J • I mm o D CLASS TUHKC
                  8240A - 41
Revision 1
 July  1992

-------
                               METHOD 8240A
VOLATILE  ORGANICS BY  GAS CHROMATOGRAPHY/MASS SPECTROMETRY (GC/MS)
                  Purg«-and-trap
                                     Dir«ct

                                     In)«cI ion
                           Selvcl
                         procedure for
                          in t roducing
                          •ample  into
                           CC/MS
                          '.2 1 S.t
                            CC/MS
                          opvnting
                         cortdi 11 oni
                        724 Connvct
                       purg»-ind-trap
                        d«vic« to CC
                        '  I 6 P.rform
                       purgcand- trap
                          analytii
' 2 8
Caleulat* RF«
for S SPCC.


7 3 P.rfor.
daily
calibration
u«ing SPCCt
and CCC.



                                 8240A  - 42
Revision  1
 July 1992

-------
                                               METHOD 8240A
                                                (continued)
          Low concentra*. ian
          sail /sediment
Medium concentration
•oil/sediment
   ?421
Dilute  sample
a', lea si  50%
 with  water
                Hater and Ma ter
                miscible liquids
                            7411
                         Screen  sampl•
                         us ing Method
                         3810  or  3820
                            7417
                            Perform
                           secondary
                          di1ution»
                          7418  Add
                       internal  standard
                         and  surrogate
                      spilt ing solution*
                           7  4  1  10
                           Per form
                        purge-and•t rap
                          procedure
                       I      '      I
                74311
             Choo»* sample
             size based on
                es tima ted
            concen t ra tion
              7 4 3 1 3 Add
            internal standard
              and surrogate
           spiking solutions
                74315
                Determine
              percent dry
                weight of
                sample
                74317
                Perform
             purge-and-trap
              procedure
                                                             7  4  1  11
                                                            Attach  trap
                                                             *.o CC  and
                                                              perform
                                                             analyst!
                                      7432 Choose
                                        sol ven t for
                                       •itraction or
                                      dilution  Weigh
                                          sample
                    7511 Identify
                      anaiytes by
                     comparing the
                   sample and standard
                     mass spectra
 7 4 3 2 2  Add
   jo 1 vent.
    shake
 7522 Calculate
 lh« concent rat von
of each identified
     analyle
   74327
    Per form
purge•and * t rap
  procedure
      7525
    Report  all
     resu1ts
                         Stop
                                                  8240A  -  43
                                                            Revision  1
                                                             July  1992

-------
00
o
03

-------
                                 METHOD 8240B

  VOLATILE ORGANIC COMPOUNDS BY GAS CHRQMATOGRAPHY/MASS SPECTROMETRY (GC/MS1
1.0   SCOPE AND APPLICATION

      1.1   Method 8240  is  used  to determine volatile organic  compounds  in a
variety of solid waste matrices.   This method  is applicable to nearly all types
of samples, regardless of water content, including ground water, aqueous sludges,
caustic  liquors,  acid liquors,  waste  solvents,  oily wastes,
fibrous  wastes,  polymeric  emulsions,   filter  cakes,  spent
catalysts, soils, and sediments.   The following compounds can
this method:
 mousses,  tars,
 carbons,  spent
be determined by
ApproDriate Technique
Analyte
Acetone
Acetonitrile
Acrolein (Propenal)
Acrylonitrile
Allyl alcohol
Allyl chloride
Benzene
Benzyl chloride
Bromoacetone
Bromochloromethane (I.S.)
Bromodichloromethane
4-Bromofluorobenzene (surr.)
Bromoform
Bromomethane
2-Butanone (MEK)
Carbon disulfide
Carbon tetrachloride
Chloral hydrate
Chlorobenzene
Chlorobenzene-d5 (I.S.)
Chi orodi bromomethane
Chloroethane
2-Chloroethanol
bis-(2-Chloroethyl) sulfide
2-Chloroethyl vinyl ether
Chloroform
Chloromethane
Chloroprene
3-Chloropropionitrile
1 , 2-Dibromo-3-chl oropropane
1,2-Dibromoethane
CAS No.b
67-64-1
75-05-8
107-02-8
107-13-1
107-18-6
107-05-1
71-43-2
100-44-7
598-31-2
74-97-5
75-27-4
460-00-4
75-25-2
74-83-9
78-93-3
75-15-0
56-23-5
302-17-0
108-90-7
3114-55-4
124-48-1
75-00-3
107-07-3
505-60-2
110-75-8
67-66-3
74-87-3
126-99-8
542-76-7
96-12-8
106-93-4
Purge-and-Trap
PP
PP
PP
PP
PP
a
a
PP
PP
a
a
a
a
a
PP
PP
a
PP
a
a
a
a
PP
PP
a
a
a
a
ND
PP
a
Direct
Injection
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
pc
pc
a
a
                                   8240B  -  1
      Revision 2
  September 1994

-------
                Appropriate  Technique
Analyte
Dibromomethane
l,4-Dichloro-2-butene
Di chl orodi f 1 uoromethane
1,1-Dichloroethane
1,2-Dichloroethane
l,2-Dichloroethane-d4(surr.)
1,1-Dichloroethene
trans -1 ,2-Dichloroethene
1,2-Dichloropropane
l,3-Dichloro-2-propanol
cis-l,3-Dichloropropene
trans-l,3-Dichloropropene
1,2,3,4-Diepoxybutane
1,4-Difluorobenzene (I.S.)
1,4-Dioxane
Epichlorohydrin
Ethanol
Ethyl benzene
Ethylene oxide
Ethyl methacrylate
2-Hexanone
2-Hydroxypropionitrile
lodomethane
Isobutyl alcohol
Malononitrile
Methacrylonitrile
Methylene chloride
Methyl iodide
Methyl methacrylate
4-Methyl -2-pentanone
Pentachloroethane
2-Picol ine
Propargyl alcohol
B-Propiolactone
Propionitrile
n-Propylamine
Pyridine
Styrene
1,1,1, 2 -Tetrachl oroethane
1,1,2 , 2-Tetrachl oroethane
Tetrachloroethene
Toluene
Toluene-d8 (surr.)
1,1,1-Trichloroethane
1, 1, 2 -Tri chl oroethane
Trichloroethene
Trichlorofl uoromethane
CAS No.b
74-95-3
764-41-0
75-71-8
75-34-3
107-06-2
107-06-2
75-35-4
156-60-5
78-87-5
96-23-1
10061-01-5
10061-02-6
1464-53-5
540-36-3
123-91-1
106-89-8
64-17-5
100-41-4
75-21-8
97-63-2
591-78-6
78-97-7
74-88-4
78-83-1
109-77-3
126-98-7
75-09-2
74-88-4
80-62-6
108-10-1
76-01-7
109-06-8
107-19-7
57-57-8
107-12-0
107-10-8
110-86-1
100-42-5
630-20-6
79-34-5
127-18-4
108-88-3
2037-26-5
71-55-6
79-00-5
79-01-6
75-69-4
Purge-and-Trap
a
PP
a
a
a
a
a
a
a
PP
a
a
a
a
PP
i
i
a
PP
a
PP
ND
a
PP
PP
PP
a
a
a
PP
i
PP
PP
PP
PP
a
i
a
a
a
a
a
a
a
a
a
a
Direct
Injection
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
pc
a
a
a
a
a
a
a
a
pc
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
8240B - 2
    Revision 2
September 1994

-------
                                                  Appropriate Technique
                                                                  Direct
Analyte                              CAS No.b   Purge-and-Trap    Injection
1,2,3-Trichloropropane
Vinyl acetate
Vinyl chloride
Xylene (Total)
96-18-4
108-05-4
75-01-4
1330-20-7
a
a
a
a
a
a
a
a
a     Adequate response by this technique.
b     Chemical Abstract Services Registry Number.
pp    Poor purging efficiency resulting in high EQLs.
i     Inappropriate technique for this analyte.
pc    Poor chromatographic behavior.
surr  Surrogate
I.S.  Internal Standard
ND    Not determined

      1.2   Method  8240  can  be  used  to  quantitate  most  volatile  organic
compounds that have boiling points below 200°C and that are  insoluble or slightly
soluble  in water.   Volatile water-soluble  compounds can  be included  in this
analytical technique.   However,  for the more  soluble compounds,  quantitation
limits are approximately  ten times  higher  because  of poor purging efficiency.
The method is  also  limited to compounds that  elute as sharp peaks  from a GC
column packed  with  graphitized carbon lightly  coated with a carbowax.   Such
compounds  include  low  molecular  weight  halogenated  hydrocarbons,  aromatics,
ketones,  nitriles,  acetates,  acrylates, ethers, and sulfides.   See Table 1 for
a list of compounds,  retention times,  and  their characteristic  ions  that have
been evaluated on a purge-and-trap GC/MS system.

      1.3   The  estimated quantitation  limit  (EQL)  of  Method  8240   for  an
individual compound  is approximately 5 /jg/kg   (wet  weight)  for soil/sediment
samples,  0.5 mg/kg  (wet weight)  for wastes,  and 5  jug/L for  ground  water (see
Table 2).  EQLs will be proportionately higher for  sample  extracts and samples
that require dilution  or reduced sample size to avoid saturation of the detector.

      1.4   This method is restricted to  use  by,  or under the  supervision of,
analysts  experienced  in   the   use  of   purge-and-trap  systems  and  gas
chromatograph/mass  spectrometers,  and skilled  in  the  interpretation  of mass
spectra and their use as a quantitative tool.

      1.5   To  increase purging  efficiencies  of acrylonitrile  and  acrolein,
refer to Methods 5030 and 8030 for proper purge-and-trap conditions.


2.0  SUMMARY OF METHOD

      2.1   The volatile compounds are introduced into the gas chromatograph by
the purge-and-trap method  or  by direct injection (in limited applications).  The


                                  8240B  -  3                         Revision 2
                                                                September 1994

-------
components are separated  via  the gas chromatograph and detected  using  a mass
spectrometer,  which  is  used  to provide  both  qualitative and  quantitative
information.     The  chromatographic  conditions,  as  well  as  typical  mass
spectrometer operating parameters, are given.

      2.2    If the above  sample  introduction techniques  are not applicable, a
portion of the sample is dispersed in methanol to dissolve the volatile organic
constituents. A portion of the methanolic solution is combined with organic-free
reagent water in a specially designed purging chamber.  It is then analyzed by
purge-and-trap GC/MS  following the normal  water method.

      2.3    The purge-and-trap process  -  An inert gas is  bubbled through the
solution at  ambient  temperature,  and the volatile  components  are efficiently
transferred  from  the aqueous phase  to  the  vapor phase.   The  vapor  is swept
through  a  sorbent column  where  the volatile components  are trapped.   After
purging is  completed, the sorbent  column  is heated and backflushed with inert gas
to  desorb  the components  onto a  gas  chromatographic  column.    The  gas
chromatographic column  is heated to elute the components, which are  detected with
a mass spectrometer.


3.0   INTERFERENCES

      3.1    Interferences  purged or  coextracted from the  samples will  vary
considerably  from  source  to  source, depending  upon the particular  sample or
extract  being  tested.   The analytical  system,  however,  should be  checked to
ensure freedom from interferences, under the analysis conditions,  by analyzing
method blanks.

      3.2    Samples  can  be  contaminated by diffusion  of  volatile  organics
(particularly methylene chloride  and fluorocarbons) through the septum seal into
the sample during  shipment and storage.   A trip blank, prepared from organic-free
reagent water and  carried  through the sampling and handling protocol, can serve
as a check on such contamination.

      3.3    Cross contamination can occur whenever high-concentration and low-
concentration  samples  are  analyzed  sequentially.    Whenever  an  unusually
concentrated  sample  is analyzed,  it should  be followed  by the  analysis  of
organic-free reagent  water to  check for cross contamination.  The purge-and-trap
system may require extensive  bake-out and cleaning  after  a high-concentration
sample.

      3.4    The  laboratory where volatile   analysis   is  performed should  be
completely free of solvents.

      3.5    Impurities in the purge gas and  from organic  compounds out-gassing
from the plumbing ahead of the trap  account  for the majority of contamination
problems.    The   analytical   system  must be  demonstrated  to  be  free  from
contamination under the conditions of the analysis  by running  calibration and
reagent blanks.  The  use of non-TFE plastic coating,  non-TFE thread sealants, or
flow controllers with rubber components  in the purging  device  should be avoided.
                                   8240B  -  4                         Revision 2
                                                                September 1994

-------
4.0   APPARATUS AND MATERIALS

      4.1   Microsyringes - 10 /xL, 25 txL, 100 /xL,  250 /xL,  500 /xL, and 1,000 /xL.
These syringes should be equipped with a 20 gauge (0.006  in. ID) needle having
a length sufficient to extend from the sample inlet  to within 1  cm of the glass
frit in the purging device.  The needle length will  depend upon the dimensions
of the purging device employed.

      4.2   Syringe valve - Two-way, with Luer ends  (three each), if applicable
to the purging device.

      4.3   Syringe - 5 ml, gas-tight with shutoff  valve.

      4.4   Balances  - Analytical,  0.0001 g,  and  top-loading, 0.1 g.

      4.5   Glass scintillation vials - 20 ml, with screw caps and Teflon liners
or glass culture tubes with a screw cap and Teflon liner.

      4.6   Volumetric flasks,  Class A - 10 ml and 100  ml,  with  ground-glass
stoppers.

      4.7   Vials  - 2 ml, for GC autosampler.

      4.8   Spatula - Stainless steel.

      4.9   Disposable pipets - Pasteur.

      4.10  Heater or  heated  oil  bath -  Should be  capable  of  maintaining the
purging chamber to within 1°C  over  the temperature range  of  ambient  to  100°C.

      4.11  Purge-and-trap device - The purge-and-trap device consists of three
separate pieces of equipment:  the  sample purger,  the trap,  and the desorber.
Several  complete devices  are commercially available.

            4.11.1     The  recommended purging chamber  is  designed to accept
      5 ml  samples with  a  water  column at  least  3  cm deep.   The  gaseous
      headspace between the water column  and the trap  must have a  total volume
      of less than 15 ml.  The purge gas  must pass through the  water column as
      finely divided bubbles with  a  diameter  of less than 3  mm  at  the  origin.
      The purge gas must be introduced no more than 5 mm from  the  base of the
      water column.   The  sample purger,  illustrated in  Figure  1,  meets  these
      design criteria.  Alternate sample purge devices may  be utilized, provided
      equivalent performance is demonstrated.

            4.11.2     The  trap must  be  at least  25 cm long and have an inside
      diameter of at least 0.105 in.  Starting from  the  inlet,  the  trap should
      contain the following  amounts of adsorbents:  1/3 of 2,6-diphenylene oxide
      polymer,  1/3 of  silica  gel,  and 1/3  of  coconut  charcoal.    It  is
      recommended that 1.0 cm of methyl  silicone coated  packing  be  inserted at
      the inlet to extend the  life of the  trap (see Figure 2).  If it is not
      necessary to analyze  for  dichlorodifluoromethane or other fluorocarbons
      of similar  volatility,  the  charcoal  can  be eliminated and the  polymer
      increased to fill 2/3 of the trap.   If only compounds boiling above 35°C


                                   8240B  - 5                         Revision 2
                                                                September 1994

-------
are to  be  analyzed,  both the silica gel  and  charcoal  can be eliminated
and the polymer  increased  to  fill  the  entire  trap.   Before initial use,
the trap should be conditioned overnight at 180°C by backflushing with an
inert gas flow of at  least  20 mL/min.  Vent the trap effluent to the room,
not to  the analytical  column.   Prior  to daily use,  the  trap should be
conditioned for  10 minutes at 180°C with backflushing.   The  trap may be
vented to the analytical column during daily conditioning.  However, the
column must be run through the  temperature program prior to  analysis of
samples.

      4.11.3      The desorber should be  capable of  rapidly  heating the
trap to 180°C  for desorption.  The  polymer section of the trap should not
be heated higher than 180°C, and  the remaining sections should not exceed
220°C  during bake out mode.  The desorber design illustrated  in Figure 2
meets these criteria.

      4.11.4      The purge-and-trap device may be assembled as a separate
unit or  may be  coupled  to a gas  chromatograph, as  shown in  Figures  3
and 4.

      4.11.5      Trap  Packing Materials

            4.11.5.1    2,6-Diphenylene  oxide  polymer  -  60/80  mesh,
      chromatographic grade (Tenax GC or  equivalent).

            4.11.5.2    Methyl   silicone  packing   -    OV-1   (3%)   on
      Chromosorb-W, 60/80  mesh or equivalent.

            4.11.5.3    Silica gel  -  35/60 mesh, Davison, grade  15  or
      equivalent.

            4.11.5.4    Coconut  charcoal - Prepare  from Barnebey Cheney,
      CA-580-26,  lot #M-2649,  by  crushing through  26  mesh  screen  (or
      equivalent).

4.12  Gas chromatograph/mass spectrometer system

      4.12.1      Gas chromatograph - An analytical  system complete with
a temperature programmable  gas chromatograph and all  required  accessories
including syringes, analytical columns, and gases.

      4.12.2      Column -  6 ft x 0.1 in. ID glass, packed with 1% SP-1000
on Carbopack-B (60/80 mesh) or equivalent.

      4.12.3      Mass  spectrometer - Capable of scanning from 35-260 amu
every 3 seconds or less, using 70 volts (nominal) electron energy in the
electron impact  mode and producing a  mass  spectrum  that  meets  all  the
criteria in Table 3 when 50 ng of 4-bromofluorobenzene (BFB) are injected
through the gas chromatograph inlet.

      4.12.4      GC/MS  interface  -  Any  GC-to-MS  interface   that  gives
acceptable calibration points at 50 ng  or less per  injection  for each of
the  analytes   and  achieves  all   acceptable  performance  criteria  (see


                            8240B  - 6                         Revision  2
                                                          September 1994

-------
      Table 3) may  be  used.  GC-to-MS  interfaces constructed entirely of glass
      or of glass-lined materials are recommended. Glass can be deactivated by
      silanizing with dichlorodimethylsilane.

            4.12.5      Data  system  -   A  computer   system  that  allows  the
      continuous acquisition and storage on machine readable media of all mass
      spectra obtained  throughout the  duration  of the chromatographic program
      must  be interfaced  to the mass  spectrometer.  The  computer  must have
      software that allows searching any GC/MS data file for ions of a specified
      mass and plotting such ion abundances versus time or  scan  number. This
      type  of plot  is  defined  as  an  Extracted  Ion  Current  Profile  (EICP).
      Software must also be available that allows  integrating the abundances in
      any EICP between  specified time  or scan  number  limits.  The  most recent
      version of the EPA/NIST Mass  Spectral Library should also be available.


5.0   REAGENTS

      5.1   Reagent  grade  chemicals  shall  be  used   in  all  tests.    Unless
otherwise indicated,  it  is  intended  that all  reagents  shall  conform  to  the
specifications of the Committee on Analytical Reagents of the American Chemical
Society, where such  specifications  are  available.  Other  grades may be used,
provided it  is first  ascertained that the reagent is of sufficiently high purity
to permit its use without lessening the accuracy of the determination.

      5.2   Organic-free  reagent water -  All  references to  water in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.3   Stock solutions - Stock  solutions may be prepared  from pure standard
materials or purchased  as  certified  solutions.  Prepare stock standard solutions
in methanol, using assayed liquids or gases, as appropriate.

            5.3.1  Place about 9.8 ml of methanol in a  10 ml tared ground-glass-
      stoppered volumetric flask.   Allow  the  flask to stand,  unstoppered,  for
      about 10 minutes or until all  alcohol wetted surfaces have dried.  Weigh
      the flask to the nearest 0.0001 g.

            5.3.2  Add  the assayed reference material,  as described below.

                   5.3.2.1    Liquids  - Using a  100 ^L syringe, immediately add
            two or more drops of assayed reference material to the flask; then
            reweigh.   The liquid must  fall directly  into  the  alcohol  without
            contacting  the neck of  the flask.

                   5.3.2.2    Gases  - To  prepare  standards for any  compounds
            that   boil   below   30°C    (e.g.    bromomethane,    chloroethane,
            chloromethane,  or  vinyl chloride), fill  a 5  ml  valved  gas-tight
            syringe with  the reference standard to the 5.0 ml mark.   Lower the
            needle to  5 mm  above  the methanol  meniscus.   Slowly introduce  the
            reference standard above the surface of the liquid.  The  heavy  gas
            will  rapidly  dissolve  in  the  methanol.   Standards  may  also  be
            prepared by using a lecture bottle equipped with a Hamilton Lecture
                                  8240B  -  7                         Revision 2
                                                                September 1994

-------
             Bottle Septum (#86600).  Attach Teflon tubing to the side-arm relief
             valve and direct a gentle stream of gas  into  the methanol meniscus.

             5.3.3 Reweigh,  dilute  to volume, stopper, and then mix  by inverting
      the flask  several  times.   Calculate the concentration in milligrams per
      liter  (mg/L) from the net gain  in weight.  When compound  purity is assayed
      to  be 96%  or greater,  the  weight  may be  used without  correction to
      calculate the concentration of the stock standard.   Commercially prepared
      stock  standards may be used at any concentration  if they  are  certified by
      the manufacturer or by an  independent  source.

             5.3.4 Transfer  the  stock  standard  solution   into  a  Teflon  sealed
      screw  cap  bottle.   Store,  with minimal headspace,   at -10°C  to -20°C and
      protect from light.

             5.3.5 Prepare  fresh  stock standards  for gases weekly or sooner if
      comparison with check standards  indicates a problem.  Reactive compounds
      such as 2-chloroethyl  vinyl ether and styrene may need to be prepared more
      frequently.  All  other standards must be replaced after  six months.  Both
      gas and liquid standards must  be monitored  closely by comparison  to the
      initial calibration curve and by  comparison to QC check standards.  It may
      be  necessary  to  replace the  standards more frequently  if  either check
      exceeds a 20% drift.

             5.3.6 Optionally, calibration using a  certified gaseous mixture can
      be  accomplished  daily utilizing  commercially  available gaseous  analyte
      mixture of bromomethane,  chloromethane,  chloroethane,   vinyl  chloride,
      dichlorodifluoromethane  and  trichlorofluoromethane  in   nitrogen.  These
      mixtures  of documented  quality  are stable  for as  long as  six  months
      without refrigeration.  (VOA-CYL  III,  RESTEK Corporation, Cat.  #20194 or
      equivalent).

      5.4    Secondary  dilution  standards  -   Using  stock  standard solutions,
prepare in methanol, secondary dilution  standards  containing  the compounds of
interest,  either singly or mixed  together.  Secondary dilution  standards must be
stored with  minimal  headspace and  should  be checked  frequently for  signs of
degradation  or  evaporation,  especially just  prior  to  preparing  calibration
standards  from them.

      5.5    Surrogate standards  -  The  surrogates  recommended  are  toluene-d8,
 4-bromofluorobenzene,  and l,2-dichloroethane-d4.   Other  compounds  may be used
as surrogates,  depending upon the  analysis  requirements.  A  stock  surrogate
solution in  methanol should be prepared as described in Sec. 5.3, and a surrogate
standard spiking solution should be prepared from the stock at a concentration
of 250 /itg/10 ml in methanol.  Each water sample undergoing GC/MS analysis must
be spiked  with 10 /A of  the surrogate  spiking solution prior to analysis.

      5.6    Internal  standards  -  The  recommended   internal  standards  are
bromochloromethane,  1,4-difluorobenzene, and  chlorobenzene-d5.  Other compounds
may be used as internal   standards as long as they have retention times  similar
to the compounds being detected by GC/MS.  Prepare internal standard stock and
secondary  dilution standards in methanol using the  procedures described in Sees.
5.3 and 5.4.  It is  recommended that the secondary dilution standard should be


                                   8240B -  8                         Revision 2
                                                                September 1994

-------
prepared  at  a concentration  of  25 mg/L  of each  internal  standard compound.
Addition of 10 juL of this standard to 5.0 mL of sample or calibration standard
would be the equivalent of 50
      5.7    4-Bromof)uorobenzene (BFB) standard - A standard solution containing
25 ng/^L of  BFB in methanol should be prepared.

      5.8    Calibration standards - Calibration standards  at a minimum of five
concentrations should be prepared from the secondary dilution of stock standards
(see Sees.  5.3 and 5.4).  Prepare these solutions in  organic-free reagent water.
One of  the concentrations should be  at  a concentration near,  but  above,  the
method detection limit. The remaining  concentrations should correspond to the
expected range of concentrations found in real samples but should not exceed the
working range of the GC/MS system.  Each standard should contain each analyte for
detection  by  this method.   It is  EPA's intent  that  all  target  analytes for a
particular analysis  be included in the calibration standard(s).   However, these
target analytes may  not include the entire List of Analytes  (Sec. 1.1) for which
the method has been demonstrated.  However,  the laboratory shall  not report a
quantitative result  for a target analyte that  was not included in the calibration
standard(s).  Calibration standards must be  prepared daily.

      5.9    Matrix  spiking standards  -  Matrix  spiking  standards  should  be
prepared from volatile organic  compounds which will  be  representative of the
compounds being investigated.   The suggested compounds are  1,1-dichloroethene,
trichloroethene, chlorobenzene, toluene,  and benzene.   The standard should be
prepared  in  methanol, with  each  compound  present  at  a concentration  of
250 /ig/10.0  mL.

      5.10   Great care must be taken to maintain the  integrity of all standard
solutions.   It is  recommended  that  all  standards in  methanol be  stored  at  -10°C
to -20°C in screw  cap amber bottles with Teflon liners.

      5.11   Methanol,  CH3OH.  Pesticide quality or equivalent. Store apart from
other solvents.

      5.12   Reagent Tetraglyme - Reagent tetraglyme is defined as tetraglyme in
which interference is not  observed at the method detection limit of compounds of
interest.

             5.12.1      Tetraglyme (tetraethylene glycol dimethyl ether, Aldrich
      #17, 240-5 or  equivalent), C8H1805. Purify by treatment at reduced pressure
      in a rotary evaporator.  The tetraglyme should have a  peroxide content of
      less  than  5  ppm as indicated  by  EM  Quant Test Strips  (available  from
      Scientific Products Co., Catalog No. P1126-8 or equivalent).

             CAUTION:    Glycol ethers  are suspected  carcinogens.  All  solvent
                        handling  should be done in a hood  while  using proper
                        protective equipment  to minimize  exposure to liquid and
                        vapor.

             Peroxides  may be removed by passing the  tetraglyme through a column
      of  activated  alumina. The  tetraglyme  is  placed in  a round  bottom flask
                                   8240B -  9                        Revision 2
                                                                September 1994

-------
      equipped with a standard taper joint, and the flask is affixed  to a rotary
      evaporator. The flask is immersed  in  a water bath at 90-100°C and a vacuum
      is  maintained  at <  10  mm  Hg for  at  least  two hours using  a two stage
      mechanical pump.  The vacuum system is equipped with  an  all  glass trap,
      which  is  maintained  in  a dry ice/methanol bath.  Cool  the tetraglyme to
      ambient temperature and  add 100 mg/L of 2,6-di-tert-butyl-4-methyl-phenol
      to  prevent peroxide  formation. Store  the tetraglyme in  a tightly sealed
      screw cap bottle  in  an  area that  is not contaminated by solvent vapors.

             5.12.2      In  order to demonstrate that all interfering volatiles
      have   been   removed   from  the   tetraglyme,   an   organic-free  reagent
      water/tetraglyme  blank  must be analyzed.

      5.13   Polyethylene glycol,  H(OCH2CH2)nOH.  Free of  interferences  at the
detection limit of the  analytes.


6.0   SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1    See  the  introductory material  to this  chapter,  Organic Analytes,
Sec. 4.1.
7.0   PROCEDURE

      Samples may be introduced  into the GC by either direct  injection or purge-
and-trap procedures.  Whichever procedure is used, the instrument calibration and
sample introduction must be performed by the same procedure.

      Regardless of which sample introduction procedure is employed, establish
GC/MS operating conditions using the following recommendations as guidance.

      Recommended GC/MS operating conditions:

            Electron energy:              70 volts  (nominal).
            Mass range:                   35-260 amu.
            Scan time:                    To  give   5 scans/peak,  but  not  to
                                          exceed 1  sec/scan.
            Initial column temperature:   45°C.
            Initial column holding time:  3 minutes.
            Column temperature  program:   8°C/minute.
            Final column temperature:     220°C.
            Final column holding time:    .15 minutes.
            Injector temperature:         200-225°C.
            Source temperature:           According    to    manufacturer's
                                          specifications.
            Transfer line temperature:    250-300°C.
            Carrier gas:                  Hydrogen at 50 cm/sec or helium at 30
                                          cm/sec.

      7.1   Direct  injection -  In  very limited  applications   (e.g.  aqueous
process wastes), direct injection of  the sample  into the GC/MS system with a 10
//L syringe may be appropriate.   One such application is for verification of the


                                  8240B - 10                        Revision 2
                                                                September 1994

-------
alcohol content  of an aqueous  sample prior  to  determining if the  sample is
ignitable (Methods 1010  or  1020).   In this case,  it  is  suggested that direct
injection be used.  The detection limit is very  high  (approximately 10,000 /^g/L);
therefore, it is  only permitted when concentrations in  excess of 10,000 /zg/L are
expected or for water soluble compounds that do not purge.  The system must be
calibrated by direct  injection using  the  procedures described in Sec. 7.2,, but
bypassing the purge-and-trap device.

      7.2    Initial calibration for  purge-and-trap procedure

             7.2.1  Establish   the  GC/MS   operating   conditions,    using   the
      recommendations in Sec. 7.0 as guidance.

             7.2.2  Each GC/MS  system must be hardware tuned to meet the criteria
      in Table 3  for  a  50 ng  injection or purging of 4-bromofluorobenzene (2 juL
      injection  of the BFB  standard).   Analyses  must  not  begin  until  these
      criteria are met.

             7.2.3  Assemble a purge-and-trap device that meets the specification
      in Sec.  4.11.   Condition the trap overnight at  180°C in the  purge  mode
      with an inert gas flow  of at least 20 mL/min.  Prior  to use,  condition the
      trap daily  for  10 min while backflushing  at  180°C with the  column at 220°C.

             7.2.4  Connect the purge-and-trap  device to a  gas chromatograph.

             7.2.5  Prepare   the  final   solutions   containing  the   required
      concentrations of  calibration  standards,  including  surrogate standards,
      directly in the purging device  (use freshly prepared stock solutions  when
      preparing  the calibration  standards  for the initial  calibration.)   Add
      5.0 ml of organic-free reagent  water  to  the purging device.   The organic-
      free  reagent water is  added to  the  purging device using  a 5 mL glass
      syringe  fitted with  a  15 cm,  20  gauge  needle.   The  needle  is  inserted
      through the sample  inlet shown  in  Figure 1.  The internal  diameter of the
      14 gauge needle that forms the  sample inlet will  permit  insertion of the
      20 gauge needle.  Next,  using a 10 pi or 25 yul microsyringe equipped  with
      a long needle (Sec. 4.1), take a volume of the secondary dilution solution
      containing  appropriate concentrations of the calibration  standards (Sec.
      5.6).  Add the aliquot  of calibration solution  directly  to  the  organic-
      free reagent water  in the purging device by inserting  the needle  through
      the sample  inlet.   When discharging the  contents of the  microsyringe, be
      sure that the end of the syringe needle is well beneath the surface of the
      organic-free reagent water.  Similarly, add  10 /uL of  the internal  standard
      solution (Sec.  5.4).   Close the 2 way syringe valve at the  sample inlet.

             7.2.6  Carry out the purge-and-trap analysis procedure  as described
      in Sec. 7.4.1.

             7.2.7  Tabulate the  area  response  of the characteristic  ions  (see
      Table  1)  against  concentration  for each  compound  and  each  internal
      standard.  Calculate response factors (RF)  for  each compound relative to
      one of the internal  standards.   The internal standard selected  for the
      calculation of the  RF  for a compound  should be the internal  standard  that
                                  8240B - 11                         Revision  2
                                                                September 1994

-------
has a retention time closest to  the  compound  being measured (Sec. 7.5.2).
The RF is calculated as follows:

      RF -  (AxCis)/(AisCx)

where:

      Ax     =     Area  of the characteristic  ion  for the compound being
                  measured.
      Ais     =     Area  of the characteristic ion for  the specific internal
                  standard.
      Cis     =     Concentration  of the specific internal  standard.
      Cx     =     Concentration  of the compound being measured.
      7.2.8 The average RF must be calculated for  each compound using the
5  RF  values  calculated  for  each compound  from  the  initial  (5-point)
calibration curve.  A system performance check should be made before this
calibration curve is used.  Five compounds (the System Performance Check
Compounds, or SPCCs) are checked for a minimum average relative response
factor.  These compounds are chloromethane,  1,1-dichloroethane, bromoform,
1,1,2,2-tetrachloroethane,  and chlorobenzene.   The minimum  acceptable
average  RF  for  these compounds  should  be 0.300  (>0.10  for bromoform).
These  compounds  typically  have  RFs of 0.4-0.6  and  are  used  to check
compound instability and to check for degradation caused by contaminated
lines or active sites in the system.  Examples of these occurrences are:

            7.2.8.1     Chloromethane -  This compound is the most likely
      compound to be lost  if  the  purge  flow  is  too  fast.

            7.2.8.2     Bromoform - This compound  is one of the compounds
      most likely to be purged very poorly if the  purge flow is too slow.
      Cold spots  and/or active sites in the transfer lines may adversely
      affect  response.   Response  of  the quantitation  ion  (m/z  173)  is
      directly  affected  by  the  tuning  of  BFB  at  ions  m/z  174/176.
       Increasing  the m/z   174/176  relative to m/z  95  ratio  may  improve
      bromoform response.

            7.2.8.3     Tetrachloroethane and 1,1-dichloroethane  - These
      compounds are degraded by contaminated transfer lines in purge-and-
      trap systems and/or  active  sites  in  trapping  materials.

      7.2.9 Using the  RFs  from  the initial  calibration,  calculate  and
record the percent relative standard deviation (%RSD) for all  compounds.
The percent RSD is calculated as follows:

              SD
      %RSD =———- x 100
              RF
where:

      RSD   =    relative standard deviation.
       RF   =    mean  of  5 initial RFs  for a compound.
      SD   =    standard deviation of  average RFs for a  compound.


                            8240B - 12                        Revision 2
                                                          September 1994

-------
       SD =
 N  (RFj - RF)
 I  	
i=l N - 1
                                 2
             where:
                  RFj    = RF for each of the 5 calibration levels
                  N     = Number of RF values (i.e., 5)

      The percent relative standard deviation should be less than 15% for
each compound.  However,  the %RSD for each individual  Calibration Check
Compound (CCC) must be less than 30%.  Late-eluting  compounds usually have
much better agreement.  The CCCs  are:

      1,1-Dichloroethene,
      Chloroform,
      1,2-Dichloropropane,
      Toluene,
      Ethylbenzene, and
      Vinyl chloride.

            7.2.9.1     If a %RSD  greater  than 30 percent is measured for
      any CCC, then corrective  action  to  eliminate a system leak and/or
      column  reactive sites is required before reattempting calibration.

      7.2.10      Linearity -  If the %RSD  of  any compound is 15% or less,
then the  relative response factor  is assumed  to  be  constant  over  the
calibration range, and the  average  relative  response factor may be used
for quantitation  (Sec. 7.5.2.2).

            7.2.10.1    If the %RSD of any compound is greater than 15%,
      construct   calibration   curves  of  area   ratio   (A/Ais)   versus
      concentration using first  or higher order regression fit  of the five
      calibration points.  The  analyst should select the regression order
      which introduces the least  calibration error into the quantitation
      (Sec. 7.5.2.4).   The use of  calibration  curves  is  a recommended
      alternative  to  average  response factor calibration,  and  a useful
      diagnostic of standard preparation accuracy and absorption activity
      in the  chromatographic system.

      7.2.11      These   curves  are  verified each  shift  by  purging  a
performance standard.   Recalibration is required only if calibration and
on-going performance criteria cannot-be met.

7.3   Daily GC/MS calibration

      7.3.1 Prior to the analysis  of samples, inject or purge 50 ng of the
4-bromofluorobenzene standard.  The resultant mass  spectra for  the  BFB
must meet  all  of  the criteria  given  in  Table 3 before  sample  analysis
begins.   These criteria must be demonstrated each 12 hour shift.
                            8240B - 13                        Revision 2
                                                          September 1994

-------
      7.3.2 The  initial calibration curve (Sec.  7.2)  for each compound of
interest must  be checked and  verified  once every 12  hours  of analysis
time.  This is accomplished  by analyzing  a calibration standard that is
at a concentration near the midpoint concentration for the working range
of the GC/MS and checking the  SPCC (Sec. 7.3.3) and CCC (Sec. 7.3.4).

      7.3.3 System   Performance  Check  Compounds  (SPCCs)  -  A  system
performance check must be made each  12  hours.   If the SPCC criteria are
met, a comparison of relative response factors is made for all compounds.
This is  the same check that is applied during  the initial  calibration.
If the minimum relative response factors are not met, the system must be
evaluated,  and  corrective  action must  be  taken before  sample  analysis
begins.   The minimum relative response factor for volatile SPCCs is 0.300
(>0.10  for Bromoform).   Some possible problems  are standard  mixture
degradation,   injection port inlet  contamination,  contamination  at  the
front end  of  the analytical column, and  active sites in  the  column or
chromatographic system.

      7.3.4 Calibration  Check Compounds   (CCCs):    After  the  system
performance check is met, CCCs  listed in Sec. 7.2.9 are used to check the
validity of the initial calibration.
      Calculate the percent drift using the following equation:

                  c, - cc
      % Drift = 	  x 100


where:
C,
      C,  =   Calibration Check Compound standard concentration.
      Cc =  Measured concentration using selected quantitation method.

      If the percent difference for each CCC is less than 20%, the initial
calibration is assumed to be valid.   If the  criterion  is  not  met (> 20%
drift),   for  any one  CCC,  corrective  action must be  taken.    Problems
similar to those listed under SPCCs could  affect  this  criterion.   If no
source of the problem can be determined after corrective action has been
taken, a new  five  point  calibration  MUST be  generated.   This criterion
MUST be met before quantitative sample analysis begins.  If the CCCs are
not required analytes by  the  permit, then all  required analytes must meet
the 20% drift criterion.

      7.3.5 The  internal  standard  responses  and retention  times  in the
check calibration standard must be evaluated immediately after or during
data acquisition.  If the retention time for any internal standard changes
by more than 30  seconds from the last  calibration check  (12 hours), the
chromatographic system must be  inspected for malfunctions and corrections
must  be  made,  as required.    If  the  EICP  area  for  any of  the  internal
standards changes by a  factor of two (- 50% to + 100%) from the last daily
calibration check standard,  the mass  spectrometer must be inspected for
malfunctions  and  corrections  must   be  made,  as  appropriate.     When
                            8240B - 14                        Revision 2
                                                          September 1994

-------
corrections are made,  reanalysis of samples  analyzed while the system was
malfunctioning is necessary.

7.4   GC/MS analysis

      7.4.1 Water samples

            7.4.1.1     Screening of the  sample  prior  to purge-and-trap
      analysis  will   provide  guidance  on  whether sample  dilution  is
      necessary  and  will  prevent contamination  of the  purge-and-trap
      system.    Two  screening  techniques  that can  be used  are:    the
      headspace  sampler (Method  3810)  using  a  gas chromatograph  (GC)
      equipped with a photo ionization  detector  (PID)  in series with an
      electrolytic conductivity  detector  (HECD); and  extraction of the
      sample with hexadecane and  analysis of  the  extract  on  a GC with a
      FID and/or an ECD (Method 3820).

            7.4.1.2     All  samples and standard solutions must be allowed
      to warm to ambient temperature before analysis.

            7.4.1.3     Set  up the GC/MS system as  outlined in Sec. 7.2.1.

            7.4.1.4     BFB  tuning criteria and daily  GC/MS  calibration
      criteria must be met (Sec. 7.3) before analyzing samples.

            7.4.1.5     Adjust  the purge  gas  (helium)  flow rate to 25-
      40 mL/min on the purge-and-trap device.   Optimize the  flow rate to
      provide the best response  for chloromethane and bromoform, if  these
      compounds are analytes.   Excessive flow rate reduces chloromethane
      response, whereas  insufficient  flow  reduces bromoform response (see
      Sec. 7.2.8).

            7.4.1.6     Remove the  plunger from a 5 ml syringe and attach
      a closed syringe valve.   Open the sample or standard bottle,  which
      has been allowed to come to ambient temperature,  and carefully pour
      the sample  into the  syringe barrel  to just short  of  overflowing.
      Replace  the  syringe  plunger and  compress  the sample.    Open  the
      syringe valve and  vent any residual  air while adjusting the sample
      volume to 5.0 ml.   This process of  taking  an  aliquot  destroys the
      validity of the liquid  sample  for future analysis; therefore,  if
      there is only one VGA vial, the analyst should fill a second syringe
      at this  time to protect  against possible loss  of  sample integrity.
      This  second  sample  is maintained  only until such  time  when  the
      analyst  has  determined  that  the  first  sample  has been  analyzed
      properly.  Filling one 20 ml syringe  would allow the use of only one
      syringe.  If a  second  analysis  is needed from a syringe, it must be
      analyzed within 24 hours.   Care must  be  taken to prevent  air from
      leaking  into the syringe.

            7.4.1.7     The   following  procedure  is  appropriate   for
      diluting purgeable samples.   All  steps must be  performed without
      delays until the diluted sample is in a gas tight syringe.
                            8240B  -  15                         Revision  2
                                                          September  1994

-------
            7.4.1.7.1    Dilutions may be made in volumetric flasks
      (10 to 100 ml).  Select the volumetric flask that will  allow
      for the  necessary dilution.   Intermediate  dilutions may be
      necessary for extremely large  dilutions.

            7.4.1.7.2    Calculate the approximate volume of organic-
      free  reagent water  to  be added  to  the  volumetric  flask
      selected and add slightly less  than this quantity of organic-
      free reagent water to the  flask.

            7.4.1.7.3    Inject  the proper aliquot of  samples  from
      the  syringe  prepared  in  Sec.  7.4.1.6  into  the  flask.
      Aliquots of  less  than  1  ml are not  recommended.   Dilute the
      sample to the mark with organic-free reagent water.  Cap the
      flask, invert, and shake  three  times.  Repeat above procedure
      for additional dilutions.

            7.4.1.7.4    Fill  a  5  ml syringe with the diluted sample
      as in Sec. 7.4.1.6.

      7.4.1.8     Add  10.0  /iL of  surrogate  spiking  solution  (Sec.
5.5) and  10  fj,l of  internal  standard spiking solution  (Sec.  5.6)
through the valve  bore  of the  syringe;  then  close the  valve.   The
surrogate and internal  standards  may  be  mixed and  added as a single
spiking solution.  The  addition  of 10 fj.L  of the surrogate spiking
solution to  5  ml  of  sample  is  equivalent  to  a  concentration of
50 /zg/L of each surrogate standard.

      7.4.1.9     Attach the  syringe-syringe valve assembly to the
syringe valve on the purging device.  Open  the  syringe valves and
inject the sample into the purging chamber.

      7.4.1.10    Close  both   valves and  purge  the  sample  for
11.0 + 0.1 minutes at ambient temperature.

      7.4.1.11    At  the conclusion of  the  purge  time,  attach the
trap to the  chromatograph, adjust the device to the desorb mode, and
begin the gas  chromatographic  temperature program  and  GC/MS  data
acquisition. Concurrently,  introduce the trapped  materials to the
gas  chromatographic  column  by rapidly  heating  the trap  to   180°C
while backflushing the trap with  inert gas between 20 and 60 ml/min
for 4 minutes.   If  this rapid heating  requirement cannot be met, the
gas  chromatographic  column  must  be  used as  a  secondary  trap by
cooling it to 30°C (or subambient, if problems  persist) instead of
the recommended initial program temperature of 45°C.

      7.4.1.12    While the  trap is  being  desorbed into  the gas
chromatograph,  empty the purging chamber.   Wash  the  chamber with a
minimum  of  two  5  ml  flushes  of organic-free  reagent water (or
methanol followed by organic-free reagent water) to avoid carryover
of pollutant compounds into subsequent analyses.
                      8240B -  16                        Revision 2
                                                    September 1994

-------
            7.4.1.13    After  desorbing  the   sample   for  4  minutes,
      recondition the trap by returning the purge-and-trap device to the
      purge mode.  Wait  15  seconds;  then  close the syringe valve on the
      purging  device  to  begin  gas  flow  through  the  trap.   The  trap
      temperature should be maintained at  180°C.  Trap temperatures up to
      220°C may be employed;  however,  the higher temperature will  shorten
      the useful life of  the trap.   After approximately 7 minutes,  turn
      off the trap heater and open the syringe  valve to  stop the gas flow
      through the trap.  When cool, the trap is  ready for the next sample.

            7.4.1.14    If the initial analysis of  a sample or a dilution
      of  the  sample has  a  concentration of  analytes  that  exceeds  the
      initial calibration range, the  sample must be reanalyzed at a higher
      dilution.  Secondary ion quantitation is allowed  only when there are
      sample  interferences   with  the  primary  ion.   When  a sample  is
      analyzed that  has  saturated  ions from a compound, this analysis must
      be followed by a blank organic-free  reagent water  analysis.  If the
      blank analysis  is not  free  of interferences,  the  system  must  be
      decontaminated.   Sample analysis may  not resume  until  a blank can
      be analyzed that is free of  interferences.

            7.4.1.15    For matrix spike analysis,  add  10 fj,l of the matrix
      spike  solution  (Sec.   5.9)  to   the  5  ml  of  sample  to  be  purged.
      Disregarding any dilutions,  this is equivalent  to a concentration
      of 50 M9/L of each  matrix spike standard.

            7.4.1.16    All   dilutions  should   keep the  response  of  the
      major constituents  (previously  saturated peaks)  in  the upper  half
      of the linear  range of the curve.  Proceed to Sees. 7.5.1 and 7.5.2
      for qualitative and quantitative analysis.

      7.4.2 Water miscible  liquids

            7.4.2.1     Water miscible  liquids  are  analyzed  as  water
      samples after  first  diluting them at least 50  fold with organic-free
      reagent water.

            7.4.2.2     Initial  and serial dilutions can  be  prepared  by
      pipetting  2 mL  of  the sample  to  a 100  ml  volumetric flask  and
      diluting  to  volume  with  organic-free   reagent  water.    Transfer
      immediately to a 5 ml gas tight syringe.

            7.4.2.3     Alternatively,, prepare  dilutions directly in a 5
      ml syringe filled with  organic-free reagent water  by  adding at least
      20 /xL,  but not  more than  100 pi of liquid sample.   The sample  is
      ready for addition of  internal   and surrogate standards.

      7.4.3 Sediment/soil and waste  samples -  It  is  highly  recommended
that all  samples of this type be screened  prior  to  the  purge-and-trap
GC/MS analysis.  The  headspace method  (Method 3810)  or  the  hexadecane
extraction  and  screening  method   (Method  3820)  may  be  used for  this
purpose.   These samples  may contain percent  quantities of  purgeable
organics that  will  contaminate the  purge-and-trap system,  and  require


                            8240B  - 17                        Revision 2
                                                          September  1994

-------
extensive cleanup  and  instrument downtime.   Use the screening  data  to
determine whether to use the low-concentration method (0.005-1 mg/kg)  or
the high-concentration method (> 1 mg/kg).

            7.4.3.1     Low-concentration method - This  is  designed for
      samples containing individual  purgeable compounds  of <  1 mg/kg.  It
      is limited to sediment/soil samples and waste that is of a similar
      consistency (granular and  porous).   The low-concentration method  is
      based on purging a heated sediment/soil sample mixed  with organic-
      free reagent water containing  the surrogate and internal standards.
      Analyze all reagent blanks and standards under the same conditions
      as the  samples.   See Figure 5 for  an  illustration of  a low soils
      impinger.

                  7.4.3.1.1    Use   a   5   g   sample   if  the  expected
            concentration  is  <  0.1  mg/kg  or a 1 g  sample  for expected
            concentrations between  0.1 and 1 mg/kg.

                  7.4.3.1.2    The GC/MS  system should be  set up  as  in
            Sees.  7.4.1.2-7.4.1.4.   This  should be  done  prior  to the
            preparation  of the  sample to  avoid  loss of volatiles from
            standards and samples.  A heated  purge calibration curve must
            be  prepared and  used  for the quantitation  of  all  samples
            analyzed  with  the  low-concentration  method.     Follow the
            initial and  daily calibration instructions, except  for the
            addition of  a 40°C purge temperature.

                  7.4.3.1.3    Remove the plunger from a 5 ml Luerlock type
            syringe  equipped  with  a  syringe  valve  and   fill  until
            overflowing  with  organic-free reagent  water.    Replace the
            plunger and  compress  the water to  vent  trapped  air.   Adjust
            the  volume  to  5.0 ml.   Add 10 /j.1  each  of surrogate  spiking
            solution (Sec. 5.5)  and  internal  standard solution (Sec. 5.6)
            to  the  syringe  through   the valve.    (Surrogate  spiking
            solution  and   internal  standard  solution  may  be  mixed
            together.)   The  addition  of  10 /nL of the surrogate  spiking
            solution to 5 g of sediment/soil  is equivalent to 50 /ig/kg  of
            each surrogate standard.

                  7.4.3.1.4    The sample  (for volatile organics) consists
            of  the entire  contents of  the  sample  container.    Do not
            discard any  supernatant liquids.  Mix  the  contents  of the
            sample  container  with  a.  narrow  metal   spatula.   Weigh the
            amount  determined  in  Sec.  7.4.3.1.1   into  a  tared  purge
            device.  Note and record the  actual weight to the nearest 0.1
            9-

                  7.4.3.1.5    Determine the  percent  dry weight  of the
            soil/sediment sample.   This  includes waste  samples  that are
            amenable to  percent dry weight determination.   Other wastes
            should be reported on a wet-weight basis.
                            8240B -  18                        Revision  2
                                                          September 1994

-------
                  7.4.3.1.5.1  Immediately after weighing the sample
            for extraction, weigh 5-10 g of the sample into a tared
            crucible.   Determine  the % dry weight of the  sample by
            drying   overnight  at   105°C.    Allow  to  cool  in  a
            desiccator   before   re-weighing.     Concentrations  of
            individual  analytes  are  reported  relative to  the dry
            weight  of sample.

                  WARNING:     The drying  oven  should  be  contained
                              in a  hood  or  vented.   Significant
                              laboratory  contamination may result
                              from a heavily contaminated hazardous
                              waste sample.

                  % dry weight =  g  of dry  sample x 100
                                  g of sample

            7.4.3.1.6    Add the  spiked water to  the purge device,
      which contains the weighed amount of sample, and connect the
      device to the purge-and-trap  system.

            NOTE:  Prior to  the attachment  of  the purge device, the
                  procedures  in  Sees.  7.4.3.1.4 and  7.4.3.1.6 must
                  be performed rapidly and without interruption to
                  avoid loss of  volatile   organics.   These  steps
                  must  be performed  in a laboratory free of solvent
                  fumes.

            7.4.3.1.7    Heat  the  sample  to 40°C +  1°C and  purge the
      sample for 11.0  + 0.1 minute.

            7.4.3.1.8    Proceed  with  the  analysis as outlined  in
      Sees. 7.4.1.11-7.4.1.16.  Use  5 ml  of  the same  organic-free
      reagent water as in  the  reagent blank.   If saturated  peaks
      occurred or  would occur if a 1 g sample  were  analyzed, the
      high-concentration method must be followed.

            7.4.3.1.9    For low-concentration  sediment/soils  add
      1C pi of the matrix spike  solution  (Sec.  5.9)  to the 5 ml of
      organic-free   reagent  water   (Sec.   7.4.3.1.3).      The
      concentration for  a  5 g  sample would be equivalent  to  50
      /ig/kg of each matrix spike standard.

      7.4.3.2     High-concentration method -  The method is based on
extracting  the  sediment/soil  with  methanol.    A waste sample  is
either  extracted   or   diluted,  depending  on  its  solubility  in
methanol.   Wastes  (i.e.   petroleum  and   coke  wastes)  that  are
insoluble  in  methanol  are  diluted  with reagent  tetraglyme  or
possibly polyethylene glycol (PEG).  An aliquot  of  the extract is
added to organic-free  reagent water containing  internal  standards.
This is  purged at ambient temperature. All  samples with an expected
concentration of > 1.0  mg/kg should be analyzed by this method.
                      8240B  -  19                        Revision  2
                                                    September 1994

-------
     7.4.3.2.1    The sample (for volatile organics)  consists
of  the  entire contents  of the  sample container.   Do  not
discard any  supernatant  liquids.   Mix  the contents of  the
sample  container  with   a  narrow  metal  spatula.     For
sediment/soil  and  solid  wastes  that   are   insoluble   in
methanol,  weigh 4 g  (wet weight) of sample into a tared  20 ml
vial.  Use a  top  loading balance.  Note and record the actual
weight  to 0.1  gram  and  determine  the percent dry weight  of
the sample using the procedure in  Sec.  7.4.3.1.5.  For  waste
that is soluble  in  methanol,  tetraglyme,  or PEG, weigh  1  g
(wet weight)  into a.tared  scintillation vial  or culture tube
or a 10 mL volumetric flask.   (If  a vial  or tube is  used,  it
must be calibrated  prior  to  use.   Pipet  10.0 ml of solvent
into the vial and mark the bottom of the  meniscus.   Discard
this solvent.)

     7.4.3.2.2    Quickly  add  9.0  ml  of  appropriate  solvent;
then add  1.0 mL  of  the  surrogate spiking  solution to  the
vial.  Cap and shake for 2 minutes.

     NOTE: Sees.  7.4.3.2.1  and 7.4.3.2.2 must  be performed
           rapidly and without interruption to avoid loss of
           volatile organics.  These steps must be performed
           in  a  laboratory free from solvent  fumes.

     7.4.3.2.3    Pipet approximately 1  ml of  the extract  to
a  GC vial  for  storage,  using  a disposable   pipet.    The
remainder may be  disposed of.  Transfer approximately 1  ml of
appropriate  solvent  to  a  separate GC  vial for use as  the
method blank  for each set  of  samples.   These extracts may be
stored at 4°C in the dark, prior to analysis.   The  addition
of  a 100  /^L  aliquot  of  each  of these extracts  in  Sec.
7.4.3.2.6 will  give a concentration equivalent to 6,200  M9/kg
of each surrogate standard.

     7.4.3.2.4    The  GC/MS system should be  set up  as  in
Sees. 7.4.1.2-7.4.1.4.   This  should  be  done  prior to  the
addition  of  the  solvent   extract  to  organic-free   reagent
water.

     7.4.3.2.5    Table 4 can be used  to determine the volume
of solvent extract to add to the 5  mL  of organic-free reagent
water for analysis.   If  a screening  procedure  was  followed
(Method 3810  or  3820),  use the estimated concentration  to
determine the  appropriate  volume.   Otherwise,  estimate  the
concentration range  of  the sample  from  the low-concentration
analysis to determine the  appropriate volume.   If the sample
was  submitted  as a  high-concentration  sample,  start  with
100 juL.   All  dilutions must  keep  the response  of the  major
constituents  (previously saturated peaks) in the upper half
of the linear range  of the curve.
               8240B  -  20                         Revision 2
                                             September  1994

-------
                  7.4.3.2.6    Remove the plunger from  a  5.0  ml Luerlock
            type  syringe  equipped with  a  syringe  valve and  fill  until
            overflowing with organic-free  reagent  water.    Replace  the
            plunger  and compress  the water to vent  trapped air.   Adjust
            the  volume  to 4.9 ml.   Pull  the plunger back to  5.0  mL to
            allow  volume  for the  addition of the sample extract  and of
            standards.  Add  10 juL of internal  standard  solution.   Also
            add  the  volume  of  solvent  extract  determined  in  Sec.
            7.4.3.2.5 and a  volume  of  extraction  or dissolution  solvent
            to total 100  jut.  (excluding  methanol in  standards).

                  7.4.3.2.7    Attach the syringe-syringe valve assembly to
            the  syringe valve on  the purging device.   Open  the  syringe
            valve  and   inject the  organic-free reagent water/methanol
            sample into the  purging chamber.

                  7.4.3.2.8    Proceed with  the  analysis as  outlined in
            Sec.  7.4.1.11-7.4.1.16.  Analyze all  reagent  blanks  on  the
            same  instrument  as that  use  for the samples.   The standards
            and blanks should also contain 100 /A  of solvent  to simulate
            the sample conditions.

                  7.4.3.2.9    For  a matrix spike  in the high-concentration
            sediment/soil  samples,  add 8.0  ml  of  methanol,  1.0 ml  of
            surrogate spike  solution  (Sec.  5.5),  and  1.0  ml  of  matrix
            spike solution (Sec.  5.9) as  in Sec. 7.4.3.2.2.   This  results
            in a 6,200 jug/kg concentration of each matrix spike standard
            when  added  to a  4 g  sample.   Add  a 100 /jL  aliquot  of  this
            extract to 5 ml of organic-free reagent water for purging (as
            per Sec. 7.4.3.2.6).

7.5   Data interpretation

      7.5.1 Qualitative analysis

            7.5.1.1     The   qualitative   identification   of   compounds
      determined  by  this method  is  based  on  retention  time,  and  on
      comparison of the sample mass spectrum,  after background  correction,
      with  characteristic ions   in  a   reference  mass  spectrum.    The
      reference mass spectrum must be generated by  the  laboratory  using
      the conditions of this method.   The characteristic  ions from  the
      reference mass  spectrum are defined to be the three ions of greatest
      relative intensity,  or  any ions over  30% relative intensity  if less
      than three such ions occur  in -the reference  spectrum.   Compounds
      should be identified as present when the criteria below, are met.

                 7.5.1.1.1    The  intensities of the characteristic  ions
            of a compound  maximize in the same scan or within  one  scan of
            each  other.   Selection  of  a peak  by a data  system  target
            compound search  routine where  the  search  is  based  on  the
            presence of  a target chromatographic  peak containing  ions
            specific  for  the  target  compound  at  a  compound-specific
            retention time will be accepted  as meeting this criterion.
                            8240B  -  21                         Revision  2
                                                          September  1994

-------
           7.5.1.1.2    The  RRT of the sample component  is  within
      + 0.06 RRT units of the RRT of the standard component.

           7.5.1.1.3    The    relative    intensities    of    the
      characteristic  ions  agree  within   30%   of   the   relative
      intensities  of  these  ions   in   the  reference   spectrum.
      (Example:   For  an ion  with  an   abundance  of 50%  in  the
      reference spectrum, the corresponding  abundance in  a  sample
      spectrum can range between 20% and 80%.)

           7.5.1.1.4    Structural isomers that produce very similar
      mass spectra  should be  identified as  individual isomers  if
      they  have   sufficiently   different  GC   retention   times.
      Sufficient GC  resolution  is achieved  if  the  height of  the
      valley between two isomer peaks is less than  25% of  the  sum
      of the two peak heights.  Otherwise,  structural isomers  are
      identified as isomeric pairs.

           7.5.1.1.5    Identification   is   hampered  when   sample
      components are not resolved chromatographically and  produce
      mass spectra  containing  ions  contributed  by  more  than  one
      analyte.  When gas chromatographic peaks obviously  represent
      more than one sample component (i.e.,  a broadened  peak with
      shoulder(s)  or  a  valley  between  two  or  more   maxima),
      appropriate  selection  of analyte  spectra  and background
      spectra  is important.   Examination of extracted ion  current
      profiles  of  appropriate   ions  can  aid  in  the selection  of
      spectra,  and  in  qualitative  identification  of compounds.
      When analytes coelute  (i.e., only one chromatographic peak is
      apparent), the identification  criteria can  be met,  but each
      analyte spectrum will  contain extraneous ions  contributed by
      the coeluting compound.

      7.5.1.2     For  samples containing  components  not associated
with the calibration standards, a library search may  be made for  the
purpose of tentative  identification.  The necessity to perform this
type of identification will  be determined by the  type of  analyses
being conducted.   Guidelines  for making tentative  identification
are:

      (1)  Relative  intensities of  major  ions  in  the  reference
spectrum (ions > 10% of  the most abundant ion) should be present in
the sample spectrum.

      (2)  The  relative  intensities of the major ions should  agree
within + 20%.   (Example:  For an ion with  an  abundance of 50% in  the
standard spectrum, the corresponding sample  ion  abundance  must  be
between 30 and 70%).

      (3)  Molecular ions present in the reference spectrum should
be present in the sample spectrum.
                      8240B -  22                         Revision  2
                                                    September  1994

-------
       (4)   Ions  present  in  the  sample spectrum  but  not  in  the
reference  spectrum  should  be  reviewed  for  possible  background
contamination or presence of coeluting  compounds.

       (5)   Ions  present in  the reference spectrum  but  not  in  the
sample spectrum should be reviewed for possible subtraction from the
sample spectrum  because of  background  contamination  or coeluting
peaks.  Data system library reduction programs can sometimes create
these discrepancies.

      Computer generated  library  search  routines should not  use
normalization  routines  that  would  misrepresent  the  library  or
unknown spectra  when compared to  each  other.  Only  after  visual
comparison of sample with the nearest library searches will the mass
spectral     interpretation    specialist    assign   a    tentative
identification.

7.5.2 Quantitative  analysis

      7.5.2.1     When   a  compound   has   been   identified,   the
quantitation  of  that compound  will  be based  on the  integrated
abundance  from  the  EICP  of  the  primary  characteristic  ion.
Quantitation will take place using  the internal standard technique.
The internal standard used shall  be  the one nearest the retention
time of that of a given  analyte (e.g. see Table 5).

      7.5.2.2     When   linearity   exists,   as  per  Sec.   7.2.10,
calculate the concentration of each identified  analyte in the sample
as follows:

      Water

                                 (A,) (I.)
      concentration (^g/L) =
                              (AJ(RF)(V0)

where:

      Ax    =      Area  of  characteristic  ion  for compound  being
                  measured.
      Is    =      Amount  of internal  standard  injected  (ng).
      Ais    =      Area  of  characteristic  ion  for  the  internal
      	          standard.
      RF    =      Mean  relative  response  factor for compound  being
                  measured  (Sec.  7.2.8).
      V0    =      Volume   of  water   purged   (ml_),  taking   into
                  consideration  any dilutions  made.
                      8240B  -  23                        Revision  2
                                                    September 1994

-------
                   Sediment/Soil   Sludge   (on  a  dry-weight  basis)   and  Waste
             (normally on  a wet-weight  basis)
                   concentration  (M9/kg) = 	
                                           (Ais)(RF)(Vi)(Ws)(D)
            where:

                   Ax> ^s'  Ais>  Rf:»  =  Same as f°r water.
                   Vt    =      Volume of total extract (jzL) (use 10,000 /xL or a
                              factor of this when dilutions are made).

                   V,    =      Volume of extract added (/iL) for purging.
                   Ws    =      Weight of sample extracted or purged (g).
                   D     =      % dry  weight of sample/100,  or  1 for a wet-weight
                              basis.

                   7.5.2.3     Where applicable, an estimate of concentration for
            noncal ibrated components in the sample should be made.  The  formulae
            given  above  should be used with  the  following modifications: The
            areas  Ax and Ais  should be  from the  total  ion chromatograms, and the
            RF  for the  compound  should be assumed to be 1.  The concentration
            obtained  should be reported  indicating  (1)  that the  value  is an
            estimate  and  (2)  which  internal  standard  was used  to  determine
            concentration.     Use  the  nearest  internal  standard   free  of
            interferences.

                   7.5.2.4     Alternatively, the regression line fitted to the
            initial  calibration  (Sec.  7.2.10.1) may  be used for determination
            of  analyte concentration.


8.0   QUALITY CONTROL

      8.1   Each  laboratory that  uses these methods  is  required  to  operate a
formal quality control program.  The minimum requirements of this program consist
of an initial  demonstration  of laboratory  capability  and an ongoing analysis of
spiked  samples  to evaluate  and  document  data  quality.    The  laboratory must
maintain records to document  the  quality  of the data generated.   Ongoing data
quality checks are compared with  established performance criteria to determine
if the results of  analyses meet the performance characteristics of the method.
When results of sample  spikes indicate atypical method performance,  a quality
control reference sample must be  analyzed  to confirm  that  the measurements were
performed in an in-control mode of operation.

      8.2   Before processing any  samples,  the  analyst  should  demonstrate,
through the analysis of a method  blank, that interferences from the analytical
system, glassware, and reagents are under  control.  Each time a set of samples
is extracted or  there is a change  in reagents, a method blank should be processed
as a  safeguard  against chronic  laboratory contamination.   The  blank samples
should be carried  through all stages of sample  preparation and measurement.
                                  8240B - 24                        Revision 2
                                                                September 1994

-------
      8.3   The  experience  of  the  analyst  performing  GC/MS  analyses  is
invaluable to the success of the  methods.   Each day that analysis is performed,
the  daily  calibration   standard  should  be  evaluated  to  determine  if  the
chromatographic system is operating properly.  Questions  that should be asked
are:  Do  the  peaks look  normal?;  Is  the response obtained  comparable  to the
response  from previous  calibrations?   Careful   examination of the  standard
chromatogram can indicate whether the column is still useable, the injector is
leaking, the injector septum needs replacing, etc.  If any changes are made to
the system (e.g. column changed), recalibration of the system must take place.

      8.4   Required  instrument  QC is found  in the following section:

            8.4.1  The GC/MS system must be tuned  to meet the  BFB specifications
      in Sec. 7.2.2.

            8.4.2  There  must  be  an  initial  calibration of the GC/MS system as
      specified in Sec. 7.2.

            8.4.3  The GC/MS system must meet the SPCC  criteria specified in Step
      7.3.3 and the CCC criteria in Sec.  7.3.4, each  12 hours.

      8.5   To  establish  the ability  to  generate   acceptable  accuracy  and
precision, the analyst must perform the following operations.

            8.5.1  A  quality  control   (QC)   reference  sample  concentrate  is
      required  containing  each  analyte  at  a  concentration  of  10  mg/L  in
      methanol.  The QC reference sample  concentrate  may be prepared from pure
      standard materials  or purchased as  certified solutions.   If  prepared by
      the laboratory, the QC  reference  sample concentrate must  be  made  using
      stock standards prepared independently from those used for calibration.

            8.5.2  Prepare a QC  reference  sample to  contain 20  jug/L of each
      analyte by adding 200 juL of QC reference sample concentrate to 100 ml of
      water.

            8.5.3  Four 5-mL aliquots of the well  mixed QC reference sample are
      analyzed according  to the method beginning  in Sec.  7.4.1.

            8.5,4  Calculate the  average recovery (x)  in jug/L, and the standard
      deviation of the recovery  (s)  in  /j.g/1, for each analyte using  the four
      results.

            8.5.5  For  each analyte  compare  s  and x  with  the  corresponding
      acceptance criteria_for  precision  and accuracy, respectively,  found in
      Table 6.  If s  and  x  for all analytes  meet the  acceptance  criteria,  the
      system performance  is  acceptable and analysis of actual  samples can_begin.
      If any individual  s exceeds the  precision limit or any  individual x falls
      outside  the  range  for   accuracy,   then   the   system  performance  is
      unacceptable for that analyte.

            NOTE:  The large number of analytes in Table 6 present a substantial
                   probability that one or more will  fail  at  least  one  of the
                                  8240B - 25                        Revision 2
                                                                September 1994

-------
                  acceptance criteria when  all  analytes  of a given method are
                  determined.

            8.5.6 When one or more of the analytes tested fail at least one of
      the  acceptance criteria,  the  analyst  must proceed  according  to  Sec.
      8.5.6.1 or 8.5.6.2.

                  8.5.6.1     Locate  and correct the  source of the problem and
            repeat the test for all analytes beginning with Sec. 8.5.2.

                  8.5.6.2     Beginning  with Sec. 8.5.2,  repeat  the test only
            for those analytes that failed to meet criteria.  Repeated failure,
            however, will confirm a general problem with the measurement system.
            If this  occurs,  locate and  correct the source  of  the  problem and
            repeat the test for  all  compounds  of interest beginning with Sec.
            8.5.2.

      8.6   The laboratory must,  on an ongoing basis,  analyze a  method blank and
a  spiked  replicate   for  each   analytical  batch  (up  to  a  maximum  of  20
samples/batch) to assess  accuracy.  For  soil and waste samples where detectable
amounts of organics  are present,  replicate samples may be appropriate in place
of spiked replicates.  For laboratories  analyzing one to ten samples per month,
at least one spiked  sample per month is  required.

            8.6.1 The  concentration  of  the  spike  in  the  sample  should  be
      determined as follows:

                  8.6.1.1     If, as in compliance monitoring, the concentration
            of  a  specific analyte  in  the  sample is  being checked  against  a
            regulatory concentration limit, the  spike  should  be  at that limit
            or 1 to  5 times higher than  the  background concentration determined
            in Sec.  8.6.2, whichever concentration would  be larger.

                  8.6.1.2     If the concentration of a  specific  analyte in  a
            water sample is not  being  checked  against  a  specific  limit,  the
            spike  should  be  at  20  /ug/L or  1  to 5  times  higher than  the
            background   concentration   determined  in  Sec.  8.6.2,  whichever
            concentration  would  be larger.   For other  matrices,  recommended
            spiking concentration  is 10 times the EQL.

            8.6.2 Analyze  one  5-mL sample aliquot to  determine the background
      concentration  (B)  of  each  analyte.   If  necessary, prepare  a   new  QC
      reference sample concentrate (Sec. 8.5.1) appropriate for the background
      concentration  in the sample.  Spike  a  second  5-mL  sample  aliquot with
      10 p,L of the QC reference sample  concentrate and analyze it to determine
      the  concentration  after spiking  (A)  of  each  analyte.   Calculate each
      percent recovery (p) as  100(A-B)%/T,  where T is  the known  true value of
      the spike.

            8.6.3 Compare  the percent recovery  (p) for each analyte in a water
      sample with the  corresponding  QC  acceptance criteria found  in Table 6.
      These  acceptance  criteria  were calculated  to  include an  allowance for
      error  in  measurement of both  the background  and  spike  concentrations,


                                  8240B - 26                        Revision  2
                                                                September 1994

-------
      assuming a spike to background  ratio of 5:1.  This error will be accounted
      for to the extent that the  analyst's spike  to background ratio approaches
      5:1.  If spiking was performed  at  a concentration  lower than 20 M9/L, the
      analyst must use either the QC acceptance criteria presented in Table 6,
      or  optional  QC  acceptance  criteria calculated  for the  specific spike
      concentration.  To calculate optional acceptance criteria for the recovery
      of  an  analyte:  (1) Calculate  accuracy (x') using the  equation  found in
      Table  7,  substituting the  spike concentration  (T)  for C;  (2)  calculate
      overall precision  (S') using the equation in Table 7, substituting x' for
      x;  (3)  calculate the range for  recovery at the  spike concentration as
      (100x'/T) + 2.44(100S'/T)%.

             8.6.4  If  any individual  p falls outside  the designated  range for
      recovery,  that  analyte  has  failed the  acceptance  criteria.   A check
      standard containing each analyte that failed the criteria must be analyzed
      as described in Sec.  8.7.

      8.7    If any analyte  in a water sample fails the  acceptance criteria for
recovery in Sec.  8.6,  a QC reference  sample containing each analyte that failed
must be prepared and analyzed.

      NOTE:  The  frequency for the required analysis  of a  QC reference sample
             will depend upon the number of analytes being simultaneously tested,
             the  complexity of the sample matrix,  and  the performance  of the
             laboratory.   If  the  entire  list  of  analytes  in Table 6  must be
             measured  in  the  sample  in  Sec.   8.6,  the probability  that  the
             analysis of a QC reference sample will be required is  high.   In this
             case, the QC reference sample should  be routinely  analyzed with the
             spiked sample.

             8.7.1  Prepare the QC reference sample by adding 10 p,l of  the QC
      reference  sample  concentrate  (Sec. 8.5.1  or  8.6.2)  to 5  ml  of  reagent
      water.  The QC  reference sample  needs only to  contain  the analytes that
      failed criteria  in  the  test in Sec. 8.6.

             8.7.2 Analyze the QC  reference sample to determine the concentration
      measured  (A)  of each analyte.   Calculate  each  percent recovery  (ps) as
      100(A/T)%, where T  is the true value of the standard concentration.

             8.7.3 Compare the percent  recovery (pj  for each analyte  with the
      corresponding QC acceptance criteria found in Table 6.   Only analytes that
      failed the test in  Sec. 8.6 need to be compared with these criteria.  If
      the recovery of any such analyte falls outside the designated range, the
      laboratory performance  for that analyte  is  judged to  be out  of control,
      and the problem must be immediately identified  and corrected.  The result
      for that analyte in the  unspiked sample is suspect and may not be reported
      for regulatory compliance purposes.

      8.8    As part of  the  QC program  for  the  laboratory,  method accuracy for
each matrix studied must be assessed  and  records must  be maintained.  After the
analysis of five spiked samplesJof the same matrix)  as in Sec.  8.6,  calculate
the  average  percent recovery  (p)  and the standard  deviation of  the  percent
recovery (sp).   Express the accuracy  assessment as a  percent  recovery  interval


                                  8240B - 27                        Revision 2
                                                                September 1994

-------
from p  -  2sp  to  p  + 2sp.   If p = 90%  and  sp  =  10%,  for example, the accuracy
interval  is  expressed as  70-110%.   Update  the accuracy  assessment  for each
analyte  on  a regular  basis  (e.g.,  after  each  five  to  ten  new  accuracy
measurements).

      8.9   To determine acceptable accuracy and precision  limits for surrogate
standards the following procedure should be performed.

            8.9.1  For each sample analyzed,  calculate the percent recovery of
      each surrogate  in the  sample.

            8.9.2  Once a minimum  of thirty samples of the same matrix have been
      analyzed,  calculate  the  average  percent  recovery  (P)   and  standard
      deviation  of  the percent recovery  (s) for  each of  the  surrogates.

            8.9.3  For a  given matrix,  calculate the upper  and lower control
      limit for method performance for  each surrogate  standard.   This should be
      done as follows:

            Upper  Control  Limit  (UCL)  =  P  + 3s
            Lower  Control  Limit  (LCL)  =  P  - 3s

            8.9.4  For aqueous and soil matrices, these laboratory established
      surrogate  control  limits  should,  if applicable,  be  compared  with  the
      control limits  listed in Table 8.  The limits given in  Table 8 are multi-
      laboratory performance based limits  for  soil  and aqueous samples,  and
      therefore, the  single-laboratory limits established in Sec.  8.9.3 must
      fall within those given in  Table 8 for these matrices.

            8.9.5  If  recovery is not within limits, the following procedures are
      required.

            •      Check  to  be  sure  there  are  no   errors  in  calculations,
                   surrogate  solutions   and internal   standards.   Also,  check
                   instrument performance.

            •      Recalculate the data and/or reanalyze the  extract if any of
                   the above  checks reveal  a problem.

            •      Reextract  and  reanalyze  the sample  if none of the above are
                   a problem  or  flag  the  data  as  "estimated concentration".

            8.9.6  At  a minimum,  each laboratory should update  surrogate recovery
      limits on  a matrix-by-matrix basis, annually.

      8.10  It  is   recommended  that  the  laboratory  adopt additional  quality
assurance practices for use with this method.   The specific practices  that are
most productive depend upon  the needs  of the  laboratory  and  the nature  of the
samples.   Field  duplicates  may  be  analyzed  to assess  the  precision of  the
environmental  measurements.  When  doubt exists  over the  identification of a peak
on the chromatogram, confirmatory techniques such as  gas chromatography  with a
dissimilar column or a different  ionization mode  using a mass spectrometer must


                                  8240B - 28                        Revision 2
                                                                September 1994

-------
be used.  Whenever  possible,  the  laboratory should analyze standard reference
materials and participate in relevant performance evaluation studies.


9.0   METHOD PERFORMANCE

      9.1   This method was tested by 15 "laboratories using organic-free reagent
water, drinking water, surface water, and industrial wastewaters spiked at six
concentrations over the range  5-600  ng/L.   Single operator precision,  overall
precision,  and method  accuracy were  found  to  be directly  related  to  the
concentration of the analyte and essentially independent of the sample matrix.
Linear equations to describe these relationships are presented in Table 7.


10.0  REFERENCES

1.    U.S. EPA 40  CFR  Part 136, "Guidelines Establishing Test Procedures for the
      Analysis  of  Pollutants  Under  the   Clean  Water  Act,   Method  624,"
      October 26,  1984.

2.    U.S.  EPA Contract  Laboratory  Program,  Statement of  Work  for  Organic
      Analysis, July  1985, Revision.

3.    Bellar, T.A., and J.J. Lichtenberg, J. Amer. Water Works Assoc.,  66(12),
      739-744, 1974.

4.    Bellar, T.A., and J.J.  Lichtenberg,  "Semi-Automated Headspace Analysis of
      Drinking  Waters and  Industrial  Waters  for Purgeable Volatile  Organic
      Compounds,"  in Van Hall, ed.,  Measurement of Organic Pollutants in Water
      and Wastewater,  ASTM STP 686,  pp. 108-129,  1979.

5.    Budde, W.L.  and J.W. Eichelberger,  "Performance Tests for the Evaluation
      of  Computerized  Gas  Chromatography/Mass  Spectrometry  Equipment  and
      Laboratories,"  EPA-600/4-79-020,  U.S. Environmental  Protection  Agency,
      Environmental Monitoring and Support Laboratory, Cincinnati, Ohio 45268,
      April 1980.

6.    Eichelberger, J.W., L.E. Harris, and  W.L.  Budde,  "Reference Compound to
      Calibrate   Ion   Abundance   Measurement   in   Gas   Chromatography-Mass
      Spectrometry Systems," Analytical Chemistry, 47, 995-1000, 1975.

7.    "Method Detection Limit for  Methods  624 and  625," Olynyk, P., W.L. Budde,
      and J.W. Eichelberger, Unpublished report, October 1980.

8.    "Interlaboratory Method Study for EPA Method 624-Purgeables," Final Report
      for EPA Contract 68-03-3102.

9.    "Method Performance Data for Method 624,"  Memorandum  from R.  Slater and
      T.  Pressley,   U.S.   Environmental    Protection   Agency,   Environmental
      Monitoring  and  Support  Laboratory,  Cincinnati, Ohio 45268,  January  17,
      1984.
                                  8240B - 29                        Revision 2
                                                                September 1994

-------
10.    Gebhart, J.E.;  Lucas,  S.V.;  Naber,  S.J.;  Berry,  A.M.;  Danison,  T.H.;
      Burkholder, H.M.  "Validation of SW-846 Methods 8010,  8015,  and 8020"; U.S.
      Environmental  Protection  Agency,  Environmental  Monitoring and  Support
      Laboratory, Cincinnati, Old 45268, July 1987, Contract No. 68-03-1760.

11.    Lucas, S.V.; Kornfeld, R.A. "GC-MS Suitability  Testing  of RCRA Appendix
      VIII and Michigan List Analytes "; U.S. Environmental  Protection Agency,
      Environmental Monitoring  and  Support  Laboratory, Cincinnati,  OH 45268,
      February 20, 1987, Contract No. 68-03-3224.
                                  8240B - 30                        Revision 2
                                                                September 1994

-------
                                   TABLE  1.
        RETENTION TIMES AND CHARACTERISTIC IONS FOR VOLATILE COMPOUNDS
Compound
Retention
Time (minutes)
Primary Ion  Secondary Ion(s)
Ethylene oxide
Chloromethane
Di chl orodi f 1 uoromethane
Bromomethane
Vinyl chloride
Acetonitrile
Chloroethane
Methyl iodide
Methylene chloride
Carbon disulfide
Trichlorofl uoromethane
Propionitrile
Allyl chloride
1,1-Dichloroethene
Bromochloromethane (I.S.)
Allyl alcohol
trans-l,2-Dichloroethene
1,2-Dichloroethane
Propargyl alcohol
Chloroform
l,2-Dichloroethane-d4(surr)
2-Butanone
Methacrylonitrile
Dibromomethane
2-Chloroethanol
b-Propiolactone
Epichlorohydrin
1,1,1-Trichloroethane
Carbon tetrachloride
1,4-Dioxane
Isobutyl alcohol
Bromodi chloromethane
Chloroprene
l,2:3,4-Diepoxybutane
1,2-Dichloropropane
Chloral hydrate (b)
cis-l,3-Dich1oropropene
Bromoacetone
Trichloroethene
Benzene
trans-l,3-Dichloropropene
1, 1, 2 -Tri chloroethane
3-Chloropropionitrile
1,2-Dibromoethane
Pyridine
1.30
2.30
2.47
3.10
3.80
3.97
4.60
5.37
6.40
7.47
8.30
8.53
8.83
9.00
9.30
9.77
10.00
10.10
10.77
11.40
12.10
12.20
12.37
12.53
12.93
13.00
13.10
13.40
13.70
13.70
13.80
14.30
14.77
14.87
15.70
15.77
15.90
16.33
16.50
17.00
17.20
17.20
17.37
18.40
18.57
44
50
85
94
62
41
64
142
84
76
101
54
76
96
128
57
96
62
55
83
65
72
41
93
49
42
57
97
117
88
43
83
53
55
63
82
75
136
130
78
75
97
54
107
79
44, 43, 42
52, 49
85, 87, 101, 103
96, 79
64, 61
41, 40, 39
66, 49
142, 127, 141
49, 51, 86
76, 78, 44
103, 66
54, 52, 55, 40
76, 41, 39, 78
61, 98
49, 130, 51
57, 58, 39
61, 98
64, 98
55, 39, 38, 53
85, 47
102
43, 72
41, 67, 39, 52, 66
93, 174, 95, 172, 176
49, 44, 43, 51, 80
42, 43, 44
57, 49, 62, 51
99, 117
119, 121
88, 58, 43, 57
43, 41, 42, 74
85, 129
53, 88, 90, 51
55, 57, 56
62, 41
44, 84, 86, 111
77, 39
43, 136, 138, 93, 95
95, 97, 132
52, 71
77, 39
83, 85, 99
54, 49, 89, 91
107, 109, 93, 188
79, 52, 51, 50
                                  8240B - 31
                                        Revision  2
                                    September  1994

-------
                                   TABLE  1.
                                  (Continued)
Compound
Retention
Time (minutes)
Primary Ion  Secondary Ion(s)
2-Chloroethyl vinyl ether
2-Hydroxypropionitrile
1,4-Difluorobenzene (I.S.)
Malononitrile
Methyl methacrylate
Bromoform
1,1,1 , 2-Tetrachl oroethane
l,3-Dichloro-2-propanol
1,1,2 , 2-Tetrachl oroethane
Tetrachloroethene
1,2,3-Trichloropropane
l,4-Dichloro-2-butene
n-Propylamine
2-Picoline
Toluene
Ethyl methacrylate
Chlorobenzene
Pentachl oroethane8
Ethyl benzene
l,2-Dibromo-3-chloropropane
4-Bromofluorobenzene (surr.)
Benzyl chloride
Styrene
bis-(2-Chloroethyl) sulfide(b)
Acetone
Acrolein
Acrylonitrile
Chlorobenzene-d5 (I.S.)
Chlorodibromomethane
1,1-Dichloroethane
Ethanol
2-Hexanone
lodomethane
4-Methyl -2-pentanone
Toluene-d8 (surr.)
Vinyl acetate
Xylene (Total)
18.60
18.97
19.60
19.60
19.77
19.80
20.33
21.83
22.10
22.20
22.20
22.73
23.00
23.20
23.50
23.53
24.60
24.83
26.40
27.23
28.30
29.50
30.83
33.53
--

--
--
--
--
--


--
--
--
	
63
44
114
66
69
173
131
79
83
164
75
75
59
93
92
69
112
167
106
157
95
91
104
109
43
56
53
117
129
63
31
43
142
43
98
43
106
65,106
44,43,42,53
63,88
66,39,65,38
69,41,100,39
171,175,252
131,133,117,119,95
79,43,81,49
85,131,133
129,131,166
75,77,110,112,97
75,53,77,124,89
59,41,39
93,66,92,78
91,65
69,41,99,86,114
114,77
167,130,132,165,169
91
157,75,155,77
174,176
91,126,65,128
104,103,78,51,77
111, 158, 160
58
55,58
52,51
82,119
208,206
65,83
45,27,46
58,57, 100
127,141
58,57,100
70,100
86
91
a The base peak at m/e 117  was not used due to an interference at  that mass with
  a nearly coeluting  internal standard, chlorobenzene-d5.
b  Response factor judged to be too low (less than 0.02) for practical use.
(I.S.) = Internal Standard
(surr) = Surrogate
                                  8240B - 32
                                        Revision 2
                                    September 1994

-------
                        TABLE  2.
ESTIMATED QUANTITATION LIMITS (EQL) FOR VOLATILE ORGANICS
                           Estimated
                          Quantitation
                            Limits8
Ground water
Volatiles /ig/L
Acetone
Acetonitrile
Allyl chloride
Benzene
Benzyl chloride
Bromodi chl oromethane
Bromoform
Bromomethane
2-Butanone
Carbon disulfide
Carbon tetrachloride
Chlorobenzene
Chl orodi bromomethane
Chloroethane
2-Chloroethyl vinyl ether
Chloroform
Chl oromethane
Chloroprene
l,2-Dibromo-3-chloropropane
1,2-Dibromoethane
Di bromomethane
l,4-Dichloro-2-butene
Di chl orodi f 1 uoromethane
1,1-Dichloroethane
1,2-Dichloroethane
1,1 Dichloroethene
trans-l,2-Dichloroethene
1,2-Dichloropropane
cis-l,3-Dichloropropene
trans-l,3-Dichloropropene
Ethyl benzene
Ethyl methacrylate
2-Hexanone
Isobutyl alcohol
Methacrylonitrile
Methyl ene chloride
Methyl iodide
Methyl methacrylate
4-Methyl -2-pentanone
Pentachloroethane
100
100
5
5
100
5
5
10
100
100
5
5
5
10
10
5
10
5
100
5
5
100
5
5
5
5
5
5
5
5
5
5
50
100
100
5
5
5
50
10
Low Soil/Sediment6
M9/kg
100
100
5
5
100
5
5
10
100
100
5
5
5
10
10
5
10
5
100
5
5
100
5
5
5
5
5
5
5
5
5
5
50
100
100
5
5
50
50
10
                       8240B  - 33                        Revision 2
                                                     September  1994

-------
                                   TABLE 2.
                                  (Continued)
                                      Estimated
                                     Quantitation
                                       Limits"
                            Ground water        Low Soil/Sediment15
Volatiles                       M9/L                  M9A9
Propionitrile
Styrene
1,1,1 , 2-Tetrachl oroethane
1,1,2 , 2-Tetrachl oroethane
Tetrachloroethene
Toluene
1,1 , 1-Trichl oroethane
1 , 1 , 2-Tri chl oroethane
Trichloroethene
1,2,3-Trichloropropane
Vinyl acetate
Vinyl chloride
Xylene (Total)
100
5
5
5
5
5
5
5
5
5
50
10
5
100
5
5
5
5
5
5
5
5
5
50
10
5
a Sample EQLs are highly matrix dependent.  The EQLs listed herein are provided
  for guidance and may not always be achievable.

b EQLs  listed  for soil/sediment are based on  wet weight.   Normally  data are
  reported on a dry weight basis; therefore,  EQLs will be higher, based on the
  percent dry weight of each sample.
               Other Matrices                      Factor0
               Water miscible liquid waste             50
               High-concentration soil and,sludge     125
               Non-water miscible waste               500
  CEQL =  [EQL for  low soil/sediment (see Table  2)]  X [Factor found  in  this
          table].  For non-aqueous samples, the factor is on  a wet weight basis.
                                  8240B - 34                        Revision 2
                                                                September 1994

-------
                                TABLE 3.
                     BFB KEY ION ABUNDANCE CRITERIA
    Mass             Ion Abundance Criteria
    50               15 to 40% of mass 95
    75               30 to 60% of mass 95
    95               base peak, 100% relative abundance
    96               5 to 9% of mass 95
   173               less than 2% of mass 174
   174               greater than 50% of mass 95
   175               5 to 9% of mass 174
   176               greater than 95% but less than 101% of mass 174
   177               5 to 9% of mass 176
                                TABLE 4.
           QUANTITY OF METHANOL  EXTRACT REQUIRED FOR ANALYSIS
                  OF  HIGH-CONCENTRATION SOILS/SEDIMENTS
       Approximate                               Volume of
   Concentration Range                        Methanol  Extract8
      500- 10,000 Mg/kg                            100  ML
    1,000- 20,000 M9/kg                             50  /uL
    5,000-100,000 Mg/kg                             10  /zL
   25,000-500,000 Mg/kg                            100  ML of 1/50 dilution6
Calculate  appropriate  dilution  factor  for  concentrations  exceeding  this
table.

a  The volume of methanol  added  to 5 mL of water being purged should be kept
   constant. Therefore,  add to the  5 mL  syringe whatever volume  of methanol
   is necessary to maintain a volume of  100 ML added to the syringe.

b  Dilute  and  aliquot  of  the  methanol  extract  and  then take  100 ML  for
   analysis.


                               8240B - 35                         Revision  2
                                                             September 1994

-------
                                   TABLE 5.
       VOLATILE INTERNAL STANDARDS WITH CORRESPONDING ANALYTES ASSIGNED
                               FOR QUANTITATION
Bromochloromethane

Acetone
Acrolein
Acrylonitrile
Bromomethane
Carbon disulfide
Chloroethane
Chloroform
Chioromethane
Di chlorodi f1uoromethane
1,1-Dichloroethane
1,2-Dichloroethane
l,2-Dichloroethane-d4 (surrogate)
1,1-Dichloroethene
trans-l,2-Dichloroethene
lodomethane
Methylene chloride
Tri chlorof1uoromethane
Vinyl chloride
1,4-Difluorobenzene

Benzene
Bromodi chloromethane
Bromoform
2-Butanone
Carbon tetrachloride
Chlorodi bromomethane
2-Chloroethyl vinyl ether
Dibromomethane
l,4-Dichloro-2-butene
1,2-Dichloropropane
cis-l,3-Dichloropropene
trans-l,3-Dichloropropene
1,1,1-Trichloroethane
1,1,2-Trichloroethane
Trichloroethene
Vinyl acetate
                        Chlorobenzene-d,;
                        Bromofluorobenzene  (surrogate)
                        Chlorobenzene
                        Ethyl benzene
                        Ethyl  methacrylate
                        2-Hexanone
                        4-Methyl-2-pentanone
                        Styrene
                        1,1,2,2-Tetrachloroethane
                        Tetrachloroethene
                        Toluene
                        Toluene-d8  (surrogate)
                        1,2,3-Trichloropropane
                        Xylene
                                  8240B - 36
                      Revision 2
                  September 1994

-------
                                TABLE 6.
                 CALIBRATION  AND  QC  ACCEPTANCE CRITERIA8


Parameter
Benzene
Bromodi chl oromethane
Bromoform
Bromomethane
Carbon tetrachloride
Chlorobenzene
2-Chloroethyl vinyl ether
Chloroform
Chl oromethane
Di bromochl oromethane
1,2-Dichlorobenzene
1,3-Di chlorobenzene
1,4-Dichlorobenzene
1,1-Dichloroethane
1,2-Dichloroethane
1,1-Dichloroethene
trans- 1,2-Di chl oroethene
1,2-Dichloropropane
cis-l,3-Dichloropropene
trans- 1 , 3-Di chl oropropene
Ethyl benzene
Methyl ene chloride
1,1,2,2-Tetrachloroethane
Tetrachloroethene
Toluene
1 , 1 , 1 -Tr i chl oroethane
1,1,2-Trichloroethane
Trichloroethene
Trichlorofluoromethane
Vinyl chloride
Range
for Q
(M9A)
12.8-27.2
13.1-26.9
14.2-25.8
2.8-37.2
14.6-25.4
13.2-26.8
D-44.8
13.5-26.5
D-40.8
13.5-26.5
12.6-27.4
14.6-25.4
12.6-27.4
14.5-25.5
13.6-26.4
10.1-29.9
13.9-26.1
6.8-33.2
4.8-35.2
10.0-30.0
11.8-28.2
12.1-27.9
12.1-27.9
14.7-25.3
14.9-25.1
15.0-25.0
14.2-25.8
13.3-26.7
9.6-30.4
0.8-39.2
Q = Concentration measured in QC check
s = Standard deviati
x = Average recovery
p, ps = Percent recovery
D = Detected; result
Limit
for s
(M9/L)
6.9
6.4
5.4
17.9
5.2
6.3
25.9
6.1
19.8
6.1
7.1
5.5
7.1
5.1
6.0
9.1
5.7
13.8
15.8
10.4
7.5
7.4
7.4
5.0
4.8
4.6
5.5
6.6
10.0
20.0
sample,
Range
for x
(M9/L)
15.2-26.0
10.1-28.0
11.4-31.1
D-41.2
17.2-23.5
16.4-27.4
D-50.4
13.7-24.2
D-45.9
13.8-26.6
11.8-34.7
17.0-28.8
11.8-34.7
14.2-28.4
14.3-27.4
3.7-42.3
13.6-28.4
3.8-36.2
1.0-39.0
7.6-32.4
17.4-26.7
D-41.0
13.5-27.2
17.0-26.6
16.6-26.7
13.7-30.1
14.3-27.1
18.5-27.6
8.9-31.5
D-43.5
in jug/L.
Range
P>Ps
(*)
37-151
35-155
45-169
D-242
70-140
37-160
D-305
51-138
D-273
53-149
18-190
59-156
18-190
59-155
49-155
D-234
54-156
D-210
D-227
17-183
37-162
D-221
46-157
64-148
47-150
52-162
52-150
71-157
17-181
D-251

on of four recovery measurements, in M9/L.
for four recovery
measured.
measurements, in M9/L.



must be greater than zero.
Criteria from 40 CFR Part 136 for Method 624 and were calculated assuming a
QC check sample concentration  of  20 M9/L-  These criteria are based directly
upon the method performance data  in Table 7.   Where necessary, the limits for
recovery  have  been  broadened to  assure applicability  of  the limits  to
concentrations below those used to develop Table 7.
                               8240B - 37
    Revision 2
September 1994

-------
                                   TABLE 7.
         METHOD ACCURACY AND PRECISION AS  FUNCTIONS OF CONCENTRATION8
Parameter
Benzene
Bromodichloromethane
Bromoform
Bromomethane
Carbon tetrachloride
Chlorobenzene
Chloroethane
2-Chloroethylvinyl ether8
Chloroform
Chloromethane
Di bromochl oromethane
l,2-Dichlorobenzeneb
1,3-Dichlorobenzene
l,4-Dichlorobenzeneb
1,1-Dichloroethane
1,2-Dichloroethane
1,1-Dichloroethene
trans-l,2,-Dichloroethene
1 , 2-Di chl oropropane8
cis-l,3-Dichloropropene8
trans-l,3-Dichloropropenea
Ethyl benzene
Methylene chloride
1,1,2 , 2-Tetrachl oroethane
Tetrachloroethene
Toluene
1,1,1-Tri chl oroethane
1,1,2-Tri chl oroethane
Trichloroethene
Tri chl orof 1 uoromethane
Vinyl chloride
Accuracy, as
recovery, x'
(09A)
0.93C+2.00
1.03C-1.58
1.18C-2.35
l.OOC
1.10C-1.68
0.98C+2.28
1.18C+0.81
l.OOC
0.93C+0.33
1.03C-1.81
1.01C-0.03
0.94C+4.47
1.06C+1.68
0.94C+4.47
1.05C+0.36
1.02C+0.45
1.12C+0.61
1.05C+0.03
l.OOC
l.OOC
l.OOC
0.98C+2.48
0.87C+1.88
0.93C+1.76
1.06C+0.60
0.98C+2.03
1.06C+0.73
0.95C+1.71
1.04C+2.27
0.99C+0.39
l.OOC
Single analyst
precision, sr'
(M/L)
0.26X-1.74
O.lSx+0.59
0.12X+0.34
0.43x
0.12X+0.25
0.16X-0.09
0.14X+2.78
0.62X
O.lGx+0.22
0.37X+2.14
0.17X-0.18
0.22X-1.45
0.14X-0.48
0.22X-1.45
0.13x-0.05
0.17X-0.32
0.17X+1.06
0.14X+0.09
0.33x
0.38x
0.25x
0.14X+1.00
0.15X+1.07
0.16X+0.69
0.13X-0.18
0.15X-0.71
0.12X-0.15
0.14X+0.02
0.13X+0.36
0.33X-1.48
0.48x
Overall
precision,
S' (ftg/i)
0.25X-1.33
0.20X+1.13
O.Ux+1.38
0.58x
O.llx+0.37
0.26X-1.92
0.29X+1.75
0.84X
0. 18X+0.16
0.58X+0.43
0.17X+0.49
0.30X-1.20
O.lSx-0.82
O.SOx-1.20
0.16x+0.47
0.21X-0.38
0.43X-0.22
0.19X+0.17
0.45x
0.52x
0.34x
0.26X-1.72
0.32X+4.00
0.20X+0.41
0.16X-0.45
0.22X-1.71
0.21x-0.39
O.lSx+O.OO
0.12X+0.59
0.34X-0.39
0.65x
x'    =      Expected  recovery  for  one  or  more  measurements   of  a  sample
             containing  a concentration  of C,  in M9/L.
s/    =      Expected  single  analyst standard deviation  of  measurements at an
             average concentration of x,  in ng/i.
S'    =      Expected  interlaboratory  standard deviation  of measurements at an
             average concentration found of x,  in  fj.g/1.
C     =      True  value  for the concentration,  in  jug/L.
x     =      Average  recovery found for measurements of  samples  containing a
             concentration of C, in  /ig/L.
a     Estimates based upon the performance in  a single laboratory.
b     Due  to chromatographic resolution problems,  performance statements  for
      these  isomers are based upon  the  sums  of their concentrations.
                                  8240B - 38
    Revision 2
September 1994

-------
                                   TABLE 8.
      SURROGATE  SPIKE  RECOVERY  LIMITS  FOR WATER AND SOIL/SEDIMENT SAMPLES
                                 Low/High             Low/High
Surrogate Compound                Water             Soil/Sediment
4-Bromofluorobenzene             86-115               74-121
l,2-Dichloroethane-d4            76-114               70-121
Toluene-do                       88-110               81-117
                                  8240B  -  39                        Revision 2
                                                               September  1994

-------
                         FIGURE 1.
                      PURGING  CHAMBER
FOAMTIMP
                       GOT 1M IN
                       M MM 00.
                       INLET 1M IN. O.O.
EXIT 1M IN. O.O
10 MM QLASS FNT
MEDIUM POM06TTY
SAMPLE INUTT

24MAV SYNNOE VALVC

17 CM 20 OAUOE SVHNQf NC£DU

« MM 0.0. RUMCR SCPTUM

INLET 1M IN O 0.
                                                  _ 1/16 IN O 0
                                                 /^ STAINUSS STffi.
                                                    13X
                                                    MOCECULAA SIEVE
                                                    PU«OE GAS FKTCT
                                                     PWWEOAS
                                                     n.ov» CONTROL
                          8240B  -  40
                                     Revision  2
                                September 1994

-------
                FIGURE 2.
TRAP PACKINGS AND CONSTRUCTION TO  INCLUDE
    DESORB  CAPABILITY FOR METHOD 8240B
    OCTM.



  -tMMOLMi
CONSTRUCTION OCTM.
 *CMTUu»QC
                8240B  -  41
                           Revision  2
                      September  1994

-------
                             FIGURE 3.
  SCHEMATIC OF PURGE-AND-TRAP  DEVICE  -  PURGE MODE FOR METHOD 8240B
CARRIER GAS
FLOW CONTROL
PRESSURE
REGULATOR
PURGE OAS
PLOW CONTROL
13X MOLECULAR
SIEVE FILTER
UOUIO INJECTION PORTS

   l— COLUMN OVEN
   UW-,
   UUTJV-
                                                CONFIRMATORY COLUMN
                                               TO DETECTOR
                                                ANALYTICAL COLUMN
                              OPTIONAL ^PORT COLUMN
                              SELECTION VALVE
                               ^	y- TRAP INLET
                                      TRAP
                                      22*C
                                PURGING
                                DEVICE
             NOTE
             ALL LINES BETWEEN TRAP
             AND QC SHOULD K HEATED
             TO
                             8240B - 42
                              Revision 2
                          September 1994

-------
                           FIGURE 4.
SCHEMATIC  OF  PURGE-AND-TRAP DEVICE - DESORB MODE FOR METHOD 82408
OARER OAS
FlOW CONTROL
PRESSURE
REGULATOR
PURGE GAS
FLOW CONTROL
 13X MOLECULAR
 SIEVE FILTER
   LJOWO INJECTION PORTS

      p- COLUMN OVEN
                                    UUUV-
                                                CONFIRMATORY COLUMN
                                               TO DETECTOR
                                                ANALYTICAL COLUMN
                              OPTIONAL 4*ORT COLUMN
                              SELECTION VALVE
                                     /-TRAP INLET
                                     TRAP
                                     200*C
i PURGING
41 DEVICE
                                              NOTE.
                                              ALL LINES BETWEEN TRAP
                                              AND OC SHOULD BE HEATED
                                              TO UPC.
                           8240B - 43
                              Revision 2
                          September 1994

-------
                    FIGURE 5.
                LOW SOILS  IMPINGER
                                 —-1
  PURGE INLET FITTING
 SAMPLE OUTLET FITTING
3" « 6mm 0 D GLASS TUBING
                                      SEPTUM
                                        CAP
             40ml VIAL
                   8240B - 44
    Revision 2
September 1994

-------
                                         METHOD 8240B
       VOLATILE  ORGANICS BY  GAS CHROMATOGRAPHY/MASS  SPECTROMETRY (GC/MS)
             7.1
            Select
         procedure for
          introducing
          sample into
           GC/MS.
 Direct
Injection
Purge-and-trap
             7.2.1
          Set GC/MS
           operating
          conditions.
         7.2.4 Connect
         purge-and-trap
         device to GC.
         7.2.6 Perform
         purge-and-trap
           analysis.
            7.2.8
         Calculate RFs
         for 6 SPCCs.
          7.3 Perform
             daily
          calibration
         using SPCCs
          and CCCs.
                                                                        Soil/Sediment
                                                        Miscible
                                                        Liquids
and Waste
 Samples
  7.4.2.1
Dilute sample
 at least 50
  fold with
   water.
               7.4.3 Screen
               sample  using
               Method 3810
                 or 3820.
                                                                  screening
                                                                method for the
                                                                    waste
                                                                   matrix.
                                                              Water
                                                             Samples
                                                                   7.4.1.1
                                                                Screen sample
                                                                using Method
                                                                3810 or 3820.
                                                                   7.4.1.7
                                                                   Perform
                                                                  secondary
                                                                  dilutions.
                                                                 7.4.1.8 Add
                                                               internal standard
                                                                and surrogate
                                                               spiking solutions.
                                                                   7.4.1.10
                                                                   Perform
                                                                purge-and-trap
                                                                  procedure.
                                          8240B  - 45
                                                                                        Revision  2
                                                                                   September  1994

-------
                                       METHOD  8240B
                                        (continued)
    7.4.3
      Is
 concentration
  > 1 mg/Kg?
   7.4.3.1.1
 Choose sample
 size based on
   estimated
 concentration.
 7.4.3.1.3 Add
internal standard
  and surrogate
spiking solutions.
   7.4.3.1.5
   Determine
   percent dry
   weight of
    sample.
    7.4.3.1.7
     Perform
  purge-and-trap
    procedure.
7.4.3.2 Choose
  solvent for
 extraction or
dilution. Weigh
    sample.
    7.4.1.1 1
   Attach trap
   to GC and
    perform
    analysis.
 7.4.3.2.2 Add
    solvent,
     shake.
7.5.1.1  Indentify
   analytes by
 comparing the
sample retention
time and sample
 mass spectra.
   7.4.3.2.7
    Perform
 purge-and-trap
   procedure.
7.5.2.2 Calculate
the concentration
of each identified
     analyte.
                                            7.5.2.4
                                           Report all
                                            results.
                                         C   Stop   J
                                        8240B  -  46
                                                  Revision 2
                                            September  1994

-------
8

-------
                                 METHOD 8250A

SEMIVOLATILE ORGANIC COMPOUNDS BY GAS CHROMATOGRAPHY/MASS SPECTROMETRY (GC/MS)


1.0   SCOPE AND APPLICATION

      1.1   Method 8250 is used to determine the concentration of semi volatile
organic compounds in extracts prepared from all  types of solid waste matrices,
soils, and ground water.  Direct  injection  of a  sample  may  be used in limited
applications.  The following compounds can be determined by  this method:
                                            Appropriate Preparation Techniques
Compounds
CAS No"   3510
                                  8250A - 1
3520  3540/   3550  3580
      3541
Acenaphthene
Acenaphthene-d10 (I.S.)
Acenaphthylene
Acetophenone
Aldrin
4-Aminobiphenyl
Aniline
Anthracene
Aroclor - 1016 (PCB-1016)
Aroclor - 1221 (PCB-1221)
Aroclor - 1232 (PCB-1232)
Aroclor - 1242 (PCB-1242)
Aroclor - 1248 (PCB-1248)
Aroclor - 1254 (PCB-1254)
Aroclor - 1260 (PCB-1260)
Benzidine
Benzoic acid
Benz(a)anthracene
Benzo(b)fluoranthene
Benzo(k)fluoranthene
Benzo(g,h,i)perylene
Benzo(a)pyrene
Benzyl alcohol
a-BHC
0-BHC
5-BHC
7-BHC (Lindane)
Bis(2-chloroethoxy)methane
Bis(2-chloroethyl) ether
Bis(2-chloroisopropyl) ether
Bis(2-ethylhexyl) phthalate
4-Bromophenyl phenyl ether
Butyl benzyl phthalate
83-32-9

208-96-8
98-86-2
309-00-2
92-67-1
62-53-3
120-12-7
12674-11-2
11104-28-2
11141-16-5
53469-21-9
12672-29-6
11097-69-1
11096-82-5
92-87-5
65-85-0
56-55-3
205-99-2
207-08-9
191-24-2
50-32-8
100-51-6
319-84-6
319-85-7
319-86-8
58-89-9
111-91-1
111-44-4
108-60-1
117-81-7
101-55-3
85-68-7
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
CP
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
NO
X
ND
X
X
X
X
X
X
X
X
X
CP
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
ND
X
ND
ND
X
X
X
X
X
X
X
X
CP
ND
X
X
X
X
X
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
ND
X
ND
X
X
X
X
X
X
X
X
X
CP
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
CP
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
                                Revision 1
                            September 1994

-------
Appropriate Preparation Techniaues
Compounds
Chlordane (technical)
4-Chloroaniline
1-Chloronaphthalene
2-Chloronaphthalene
4-Chloro -3 -methyl phenol
2-Chlorophenol
4-Chlorophenyl phenyl ether
Chrysene
Chrysene-d12 (I.S.)
4,4'-DDD
4,4'-DDT
4,4'-DOE
Dibenz(a, j)acridine
Dibenz( a, h) anthracene
Dibenzofuran
Di-n-butyl phthalate
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
l,4-Dichlorobenzene-d4 (I.S)
3,3' -Dichlorobenzidine
2,4-Dichlorophenol
2,6-Dichlorophenol
Dieldrin
Diethyl phthalate
Dimethyl aminoazobenzene
7,12-Dimethylbenz(a)-
anthracene
a,a-Dimethylphenethylamine
2,4-Dimethylphenol
Dimethyl phthalate
4, 6-Dinitro-2-methyl phenol
2,4-Dinitrophenol
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Di phenyl ami ne
1,2-Di phenyl hydrazine
Di-n-octyl phthalate
Endosulfan I
Endosulfan II
Endosulfan sulfate
Endrin
Endrin aldehyde
Endrin ketone
Ethyl methanesulfonate
CAS Noa
57-74-9
106-47-8
90-13-1
91-58-7
59-50-7
95-57-8
7005-72-3
218-01-9

72-54-8
50-29-3
72-55-9
224-42-0
53-70-3
132-64-9
84-74-2
95-50-1
541-73-1
106-46-7
3855-82-1
91-94-1
120-83-2
87-65-0
60-57-1
84-66-2
60-11-7

57-97-6
122-09-8
105-67-9
131-11-3
534-52-1
51-28-5
121-14-2
606-20-2
122-39-4
122-66-7
117-84-0
959-98-8
33213-65-9
1031-07-8
72-20-8
7421-93-4
53494-70-5
62-50-0
3510
X
X
X
X

X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X

CP(45)
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
3520
X
ND
X
X

X
X
X
X
X
X
X
ND
X
X
X
X
X
X
X
X
X
ND
X
X
ND

ND
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
ND
3540/
3541
X
ND
X
X

X
X
X
X
X
X
X
ND
X
ND
X
X
X
X
X
X
X
ND
X
X
ND

ND
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
X
ND
ND
3550
X
ND
X
X

X
X
X
X
X
X
X
ND
X
X
X
X
X
X
X
X
X
ND
X
X
ND

ND
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
ND
3580
X
X
X
X

X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X

CP
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
8250A - 2
    Revision 1
September 1994

-------
Compounds
        Appropriate Preparation Techniques

CAS Noa   3510     3520  3540/   3550  3580
                        3541
Fluoranthene
Fluorene
2-Fluorobiphenyl (surr.)
2-Fluorophenol (surr.)
Heptachlor
Heptachlor epoxide
Hexachlorobenzene
Hexachlorobutadiene
Hexachlorocyclopentadiene
Hexachloroethane
Indeno(l,2,3-cd}pyrene
Isophorone
Methoxychlor
3-Methylchol anthrene
Methyl methanesulfonate
2 -Methyl naphthalene
2-Methylphenol
4-Methylphenol
Naphthalene
Naphthalene-d8 (I.S.)
1-Naphthylamine
2-Naphthylamine
2-Nitroaniline
3-Nitroaniline
4-Nitroaniline
Nitrobenzene
Nitrobenzene-dg (surr.)
2-Nitrophenol
4-Nitrophenol
N -Nitrosodi butyl ami ne
N-Nitrosodi methyl ami ne
N-Nitrosodiphenylamine
N-Nitrosodi-n-propylamine
N-Nitrosopi peri dine
Pentachlorobenzene
Pentachloronitrobenzene
Pentachlorophenol
Perylene-d12 (I.S.)
Phenacetin
Phenanthrene
Phenanthrene-d10 (I.S.)
Phenol
Phenol -d6 (surr.)
2-Picoline
Pronamide
206-44-0
86-73-7
321-60-8
367-12-4
76-44-8
1024-57-3
118-74-1
87-68-3
77-47-4
67-72-1
193-39-5
78-59-1
72-43-5
56-49-5
66-27-3
91-57-6
95-48-7
106-44-5
91-20-3
1146-65-2
134-32-7
91-59-8
88-74-4
99-09-2
100-01-6
98-95-3
4165-60-0
88-75-5
100-02-7
924-16-3
62-75-9
86-30-6
621-64-7
100-75-4
608-93-5
82-68-8
87-86-5
198-55-0
62-44-2
85-01-8

108-95-2
13127-88-3
109-06-8
23950-58-5
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
OS(44)
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
DC(28)
DC(28)
NO
X
X
X
X
X
X
X
X
X
X
X
X
X
ND
ND
ND
X
ND
ND
X
X
ND
ND
X
X
X
X
X
X
X
ND
X
X
X
ND
ND
ND
X
X
ND
X
X
X
X
ND
ND
X
X
X
X
X
X
X
X
X
X
X
X
ND
ND
ND
ND
ND
ND
X
X
ND
ND
ND
ND
ND
X
X
X
X
ND
X
X
X
ND
ND
ND
X
X
ND
X
X
X
X
ND
ND
X
X
X
X
X
X
X
X
X
X
X
X
ND
ND
ND
X
ND
ND
X
X
ND
ND
X
X
X
X
X
X
X
ND
X
X
X
ND
ND
ND
X
X
ND
X
X
X
X
ND
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
ND
X
                                   8250A -  3
                                Revision 1
                            September 1994

-------
                                            Appropriate Preparation Techniques

Compounds                           CAS Noa   3510     3520  3540/   3550  3580
                                                            3541
Pyrene
Terpheny1-d14(surr.)
1,2,4 , 5-Tetrachl orobenzene
2,3,4,6-Tetrachlorophenol
Toxaphene
2,4,6-Tribromophenol (surr. )
1,2, 4 -Trichl orobenzene
2,4,5-Trichlorophenol
2,4,6-Trichlorophenol
129-00-0
1718-51-0
95-94-3
58-90-2
8001-35-2
118-79-6
120-82-1
95-95-4
88-06-2
X
X
X
X
X
X
X
X
X
X
X
ND
ND
X
X
X
X
X
X
ND
ND
ND
X
X
X
ND
X
X
X
ND
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
X
8     Chemical Abstract Service Registry Number.

CP    =     Nonreproducible chromatographic performance.
DC    =     Unfavorable  distribution  coefficient  (number  in  parenthesis  is
            percent recovery).
ND    =     Not determined.
OS    =     Oxidation  during   storage   (number  in   parenthesis  is  percent
            stability).
X     =     Greater than 70 percent recovery by this technique.


      1.2   Method 8250  can be used  to quantitate most  neutral,  acidic,  and
basic organic compounds that  are  soluble  in  methylene  chloride and capable of
being eluted without derivatization  as  sharp  peaks  from a gas chromatographic
packed  column.    Such compounds  include  polynuclear   aromatic  hydrocarbons,
chlorinated  hydrocarbons   and  pesticides,  phthalate  esters,  organophosphate
esters,  nitrosamines,  haloethers,   aldehydes,   ethers,   ketones,  anilines,
pyridines,   quinolines,  aromatic  nitro  compounds,   and  phenols,  including
nitrophenols.  See Table 1  for a list of compounds  and their characteristic ions
that have been evaluated on the specified GC/MS system.

      1.3   The following  compounds  may require  special  treatment when being
determined by this method.   Benzidine can  be  subject to  oxidative losses during
solvent  concentration.   Also,  chromatography  is  poor.   Under  the alkaline
conditions of the  extraction step, a-BHC, 7-BHC, endosulfan I  and II,  and endrin
are subject to decomposition.  Neutral extraction should be performed if these
compounds  are  expected   and   are   not  being  determined  by  Method  8080.
Hexachlorocyclopentadiene is subject to thermal  decomposition in the inlet of the
gas chromatograph,  chemical  reaction  in  acetone solution,  and photochemical
decomposition.  N-nitrosodimethylamine is difficult to  separate from the solvent
under  the  chromatographic   conditions   described.     N-nitrosodiphenylamine
decomposes  in the  gas chromatographic  inlet  and  cannot  be  separated  from
diphenylamine. Pentachlorophenol,  2,4-dinitrophenol, 4-nitrophenol, 4,6-dinitro-
2-methylphenol,   4-chloro-3-methylphenol,   benzoic   acid,    2-nitroaniline,


                                  8250A -  4                         Revision 1
                                                                September 1994

-------
3-nitroaniline,  4-chloroaniline,  and  benzyl  alcohol  are subject  to  erratic
chromatographic behavior, especially if the GC  system is contaminated with high
boil ing material.

      1.4   The  estimated   quantitation   limit   (EQL)   of  Method  8250  for
determining an  individual  compound  is approximately  1 mg/kg  (wet weight) for
soil/sediment samples,  1-200 mg/kg for wastes (dependent on matrix  and method of
preparation),  and 10 /xg/L for ground water samples  (see Table  2).  EQLs will be
proportionately  higher for  sample  extracts  that   require  dilution to  avoid
saturation of the detector.

      1.5   This method  is restricted to  use  by or  under the  supervision of
analysts experienced  in  the use  of gas  chromatograph/mass  spectrometers and
skilled in  the  interpretation of mass spectra.  Each  analyst must demonstrate the
ability to generate acceptable results with this method.


2.0   SUMMARY OF METHOD

      2.1   Prior  to using  this  method,  the  samples should be  prepared for
chromatography using  the  appropriate sample preparation  and  cleanup  methods.
This  method  describes  chromatographic  conditions   that  will  allow  for  the
separation of the compounds in the extract.


3.0   INTERFERENCES

      3.1   Raw  GC/MS data from  all  blanks,   samples,  and  spikes  must  be
evaluated for interferences.  Determine if the  source  of interference is in the
preparation and/or  cleanup of the samples and take corrective action to eliminate
the problem.

      3.2   Contamination by carryover can occur whenever high-concentration and
low-concentration samples are sequentially analyzed.  To reduce carryover, the
sample syringe must  be rinsed  out between samples  with solvent.   Whenever an
unusually  concentrated  sample is encountered,  it   should  be followed by  the
analysis of solvent to check for cross contamination.


4.0   APPARATUS AND MATERIALS

      4.1   Gas  chromatograph/mass spectrometer  system

            4.1.1 Gas  chromatograph  -  An analytical system  complete with  a
      temperature-programmable  gas   chromatograph    suitable   for   splitless
      injection  and  all  required accessories,  including  syringes,  analytical
      columns,  and gases.

            4.1.2 Columns

                  4.1.2.1      For base/neutral  compound detection  -  2  m  x  2
            mm ID stainless or glass,  packed with 3% SP-2250-DB on 100/120 mesh
            Supelcoport or equivalent.


                                  8250A -  5                         Revision  1
                                                                September 1994

-------
                   4.1.2.2      For acid compound detection - 2 m x 2 mm ID glass,
            packed with 1% SP-1240-DA on 100/120 mesh Supelcoport  or equivalent.

            4.1.3  Mass  spectrometer  -  Capable of  scanning  from  35 to 500 amu
      every 1  second  or  less,  using  70 volts (nominal) electron energy in the
      electron  impact  ionization mode.   The mass spectrometer must be capable
      of  producing a  mass  spectrum for  decafluorotriphenylphosphine (DFTPP)
      which meets all of the criteria in Table 3 when  1 /uL of the  GC/MS tuning
      standard  is  injected through the GC  (50 ng of DFTPP).

            4.1.4  GC/MS  interface - Any GC-to-MS interface that gives acceptable
      calibration  points at  50  ng per  injection  for each compound of interest
      and achieves acceptable tuning  performance criteria may be used.  GC-to-MS
      interfaces  constructed entirely of  glass  or glass-lined  materials are
      recommended.      Glass   may    be    deactivated    by   silanizing   with
      di chlorodimethylsi 1ane.

            4.1.5  Data system - A computer  system must be interfaced to the mass
      spectrometer.   The  system  must  allow the  continuous acquisition  and
      storage on machine-readable  media of  all mass spectra obtained throughout
      the  duration of the  chromatographic program.   The computer  must  have
      software that can  search any GC/MS data file for ions of a specific mass
      and that  can plot  such ion  abundances versus time or scan  number.   This
      type  of plot  is  defined as  an Extracted  Ion   Current Profile (EICP).
      Software must also be available that  allows integrating the abundances in
      any EICP  between specified time  or  scan-number  limits.   The most recent
      version of the EPA/NIH Mass Spectral   Library should also be  available.

      4.2   Syringe -  10 /xL.


5.0   REAGENTS

      5.1   Reagent grade chemicals  shall be used in all tests. Unless otherwise
indicated, it is intended that all reagents shall conform to the specifications
of the Committee on Analytical  Reagents of  the American Chemical  Society,  where
such specifications are available. Other grades may  be used, provided it is first
ascertained that the reagent  is of sufficiently  high  purity to  permit its use
without lessening the accuracy of the determination.

      5.2   Organic-free reagent water.   All references to water  in this method
refer to organic-free reagent water,  as defined in Chapter One.

      5.3   Stock  standard  solutions (lOO'O mg/L)  - Standard  solutions  can be
prepared from pure standard materials or purchased as certified  solutions.

            5.3.1  Prepare stock standard solutions  by accurately weighing about
      0.0100  g  of  pure material.   Dissolve the material  in pesticide quality
      acetone  or other  suitable  solvent  and dilute  to  volume  in  a 10  ml
      volumetric flask.  Larger volumes can be used  at the convenience of the
      analyst.  When  compound purity  is assayed to be 96% or greater, the weight
      may be used without correction to calculate  the concentration  of the stock
      standard.   Commercially  prepared  stock standards may be  used at  any


                                  8250A  -  6                        Revision 1
                                                                September 1994

-------
      concentration  if  they  are   certified  by  the  manufacturer  or  by  an
      independent source.

            5.3.2 Transfer the stock standard solutions into bottles with Teflon
      lined screw-caps  or crimp tops.   Store  at  -10°C  to  -20°C  or less and
      protect from light. Stock standard solutions should be checked frequently
      for  signs  of degradation  or  evaporation,  especially  just  prior  to
      preparing calibration standards from them.

            5.3.3 Stock  standard  solutions must  be replaced after 1 year or
      sooner  if comparison  with  quality  control  check samples  indicates  a
      problem.

      5.4   Internal standard solutions  - The internal standards  recommended are
l,4-dichlorobenzene-d4,    naphthalene-d8,   acenaphthene-d10,   phenanthrene-d10,
chrysene-d12,  and perylene-d12.  Other compounds may be used as internal  standards
as long as  the requirements given in Sec. 7.3.2 are met.  Dissolve 200 mg of each
compound with  a  small volume of carbon disulfide.  Transfer to a 50 ml volumetric
flask and dilute to  volume with methylene chloride  so that the final  solvent is
approximately 20% carbon disulfide.   Most of the compounds are also  soluble in
small volumes of methanol,  acetone,  or  toluene, except  for  perylene-d12.   The
resulting solution will contain each standard at  a concentration  of 4,000 ng//xL.
Each 1 ml sample extract  undergoing  analysis should  be spiked with 10 fj,L of the
internal standard solution,  resulting in  a concentration of  40  ng/nl of each
internal standard.  Store at -10°C to -20°C or less when not being  used.

      5.5   GC/MS tuning standard  - A  methylene  chloride  solution containing
50 ng/VL  of  decafluorotriphenylphosphine  (DFTPP)  should   be  prepared.    The
standard should also contain 50 ng/^L each of 4,4'-DDT,   pentachlorophenol, and
benzidine to verify injection port inertness and GC column performance.  Store
at 4°C or less when  not  being used.

      5.6   Calibration  standards - Calibration standards at a minimum of five
concentrations should be prepared.   One of the calibration standards  should be
at a concentration near, but  above, the method detection limit; the others should
correspond to the range of concentrations found  in real   samples  but should not
exceed the working range  of the GC/MS system.   Each  standard  should contain each
analyte for detection by this method  (e.g.  some  or all of the compounds listed
in Table 1 may be included).  Each 1 mL aliquot  of calibration standard should
be spiked with 10 ^L of  the internal standard solution prior to analysis.  All
standards should be  stored at  -10°C to -20°C and  should be freshly prepared once
a year,  or sooner if check standards  indicate a  problem.  The daily  calibration
standard should be prepared  weekly and stored at 4°C.

      5.7   Surrogate  standards   -   The  recommended surrogate  standards are
phenol-d6,     2-fluorophenol,    2,4,6-tribromophenol,    nitrobenzene-d5,    2-
fluorobiphenyl,  and p-terphenyl-d14.   See Method 3500 for  the instructions on
preparing the  surrogate standards.  Determine what concentration should be in the
blank extracts after all  extraction, cleanup,  and concentration  steps.  Inject
this concentration into the  GC/MS  to determine recovery of surrogate standards
in all blanks, spikes,  and sample extracts.  Take into account all dilutions of
sample extracts.
                                  8250A  -  7                         Revision 1
                                                                September 1994

-------
      5,8   Matrix  spike  standards  -  See  Method  3500  for  instructions  on
preparing the matrix spike standard.  Determine what concentration should be in
the blank  extracts  after  all  extraction,  cleanup,  and  concentration  steps.
Inject this concentration into the GC/MS to determine recovery of standards in
all matrix spikes.  Take into account all dilutions of sample extracts.


6.0   SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1   See  the  introductory material to this  chapter,  Organic Analytes,
Sec. 4.1.
7.0   PROCEDURE

      7.1   Sample  preparation  -  Samples  must  be  prepared
following methods prior to GC/MS analysis.
                         by  one  of  the
      Matrix
      Water
      Soil/sediment
      Waste
      Methods
      3510,  3520
      3540,  3541,  3550
      3540,  3541,  3550,  3580
            7.1.1 Direct  injection  -  In  very  limited  applications  direct
      injection of the sample into  the  GC/MS  system with a 10 /xL syringe may be
      appropriate.     The  detection   limit   is  very   high   (approximately
      10,000 M9/L);  therefore,  it  is  only  permitted where  concentrations  in
      excess of  10,000 /xg/L are expected.   The  system must be  calibrated  by
      direct injection.
      7.2   Extract cleanup - Extracts
methods prior to GC/MS analysis.
may be cleaned  up by any of the  following
      Compounds
      Phenols
      Phthalate esters
      Nitrosamines
      Organochlorine pesticides & PCBs
      Nitroaromatics and cyclic ketones
      Polynuclear aromatic hydrocarbons
      Haloethers
      Chlorinated hydrocarbons
      Organophosphorus pesticides
      Petroleum waste
      All basic, neutral, and acidic
      Priority Pollutants
         Methods
         3630, 3640, 8040°
         3610, 3620, 3640
         3610, 3620, 3640
         3620, 3640, 3660
         3620, 3640
         3611, 3630, 3640
         3620, 3640
         3620, 3640
         3620
         3611, 3650

         3640
"Method 8040  includes a derivatization technique followed  by GC/ECD analysis, if
interferences are encountered on GC/FID.
                                   8250A  -  8
                              Revision  1
                          September  1994

-------
7.3   Recommended GC/MS operating conditions

Electron energy:               70 volts (nominal)
Mass range:                    35-500 amu
Scan time:                     1  sec/scan
Injector temperature:          250-300°C
Transfer line temperature:     250-300°C
Source temperature:            According  to manufacturer's specifications
Injector:                      Grob-type,  splitless
Sample volume:                 1-2 juL
Carrier gas:                   Helium at  30  mL/min

Conditions for base/neutral analysis  (3% SP-2250-DB):

Initial  column temperature and hold time:       50°C for 4 minutes
Column temperature program:                     50-300°C at 8°C/min
Final column temperature hold:                  300°C for 20 minutes

Conditions for acid analysis  (1% SP-1240-DA):

Initial  column temperature and hold time:       70°C for 2 minutes
Column temperature program:                     70-200°C at 8°C/min
Final column temperature hold:                  200°C for 20 minutes

7.4 Initial calibration

      7,4.1 Each GC/MS system must be hardware-tuned to  meet the criteria
in Table  3  for  a 50 ng injection of DFTPP.  Analyses  should  not begin
until all  these  criteria  are met.    Background  subtraction  should  be
straightforward and designed only to  eliminate column bleed or instrument
background ions.  The GC/MS tuning standard  should also  be used to assess
GC column performance and  injection  port  inertness.   Degradation of DDT
to DDE  and  ODD  should not  exceed 20% (See  Sec.  7.4.5  of Method 8080).
Benzidine  and  pentachlorophenol should be  present   at  their  normal
responses,  and  no peak tailing  should be  visible.   If  degradation  is
excessive and/or  poor  chromatography  is  noted,  the injection  port may
require cleaning.

      7.4.2 The internal standards selected  in Sec. 5.1 should permit most
of the components of interest in a  chromatogram to have retention times
of 0.80-1.20 relative to one of the internal  standards.   Use the base peak
ion  from  the  specific  internal  standard  as  the  primary  ion  for
quantitation (see Table  1).  If interferences are noted,  use the next most
intense ion as the quantitation  ion  (i.e. for l,4-dichlorobenzene-d4 use
m/z 152 for quantitation).
                            8250A  -  9                         Revision 1
                                                          September 1994

-------
      7.4.3 Analyze 1 /j,l of each calibration standard (containing  internal
standards) and tabulate the area of the primary characteristic  ion against
concentration  for  each compound  (as  indicated in  Table  1).   Calculate
response factors (RFs) for each compound relative to the internal  standard
as follows:

      RF = (AxCis)/(AisCJ

where:

Ax    =     Area  of  the  characteristic  ion  for  the compound  being
            measured.
Ais    =     Area  of  the  characteristic  ion  for  the  specific  internal
            standard.
Cx    =     Concentration  of the  compound  being measured  (ng//iL).
Cis    =     Concentration  of the  specific  internal  standard (ng/^L).

      7.4.4 A  system  performance  check must  be performed to ensure that
minimum  average response factors,  calculated  as  the mean  of  the  5
individual relative response factors, are met before the calibration curve
is  used.    For semivolatiles,  the System Performance  Check Compounds
(SPCCs)  are:     N-nitroso-di-n-propylamine;  hexachlorocyclopentadiene;
2,4-dinitrophenol;  and 4-nitrophenol.   The minimum  acceptable average RF
for these  compounds  is  0.050.   These  SPCCs typically  have  very  low RFs
(0.1-0.2) and tend to decrease in response as the chromatographic system
begins to  deteriorate or  the  standard material  begins to  deteriorate.
They are usually the first  to show poor performance.  Therefore, they must
meet the minimum requirement when the  system is calibrated.

            7.4.4.1     The percent relative standard deviation should be
      less  than 15%  for  each  compound.   However, the  %RSD for each
      individual Calibration Check  Compound  (CCC) (see Table  4)  must be
      less than 30%.   The relative retention  times of each compound in
      each calibration  run should agree within  0.06 relative retention
      time  units.     Late-eluting  compounds  usually  have  much  better
      agreement.
                        SD
            %RSD = 	—	   x 100
                        RF
      where:
            RSD   =     relative  standard deviation.
            RF    =     mean of 5  initial RFs  for  a compound.
            SD    =     standard deviation of average  RFs  for a compound.
                            8250A - 10                        Revision 1
                                                          September 1994

-------
            SD =
 N  (RFj - RF):
 I  	
1=1  N - 1
      where:
            RFi    =      RF  for each  of  the  5  calibration  levels
            N     =      Number of  RF values  (i.e.,  5)

            7.4.4.2      If the %RSD of any CCC is 30% or greater, then the
      chromatographic system is too reactive for analysis to begin.  Clean
      or replace the injector liner  and/or capillary column, then repeat
      the calibration procedure beginning with Sec. 7.4.

      7.4.5 Linearity -  If the %RSD  of any compound is 15% or less, then
the  relative  response   factor   is  assumed   to  be  constant  over  the
calibration range, and the  average relative  response  factor may be used
for quantitation (Sec.  7.7.2).

            7.4.5.1      If the %RSD  of any compound is greater than 15%,
      construct   calibration   curves   of  area  ratio   (A/Ais)   versus
      concentration using first or  higher order regression fit of the five
      calibration points.  The analyst should  select the regression order
      which introduces the least  calibration error into the quantitation
      (Sees. 7.7.2.2 and  7.7.2.3).  The use of calibration  curves  is  a
      recommended alternative to  average response factor calibration, and
      a useful  diagnostic of standard preparation accuracy and absorption
      activity in the chromatographic system.

7.5   Daily GC/MS calibration

      7.5.1 Prior to analysis of samples,  the GC/MS tuning standard must
be analyzed.  A  50 ng injection  of DFTPP  must result  in a mass spectrum
for DFTPP which meets the criteria  given in Table 3.  These criteria must
be demonstrated during each 12 hour  shift.

      7.5.2 A calibration standard(s) at mid-concentration containing all
semivolatile  analytes,   including   all  required  surrogates,   must  be
analyzed every  12 hours during analysis.   Compare the instrument response
factor from the  standards every  12 hours  with  the  SPCC (Sec. 7.5.3) and
CCC (Sec. 7.5.4) criteria.

      7.5.3 System  Performance  Check  Compounds  (SPCCs)  -  A  system
performance check must be made during  every  12  hour  shift.   If the SPCC
criteria  are  met, a comparison   of   response  factors   is  made for  all
compounds.  This is  the same check  that  is applied during  the  initial
calibration. If the minimum response factors are not met, the system must
be evaluated,  and corrective action must be  taken  before sample analysis
begins.   The minimum RF  for semivolatile  SPCCs  is  0.050.   Some possible
problems  are   standard   mixture  degradation,  injection   port   inlet
contamination,  contamination  at the  front end  of  the  analytical  column,
                            8250A - 11                        Revision 1
                                                          September 1994

-------
and active sites  in the column or chromatographic system.  This check must
be met before analysis begins.

      7.5.4 Calibration  Check  Compounds  (CCCs):    After  the  system
performance check is  met,  CCCs  listed in Table 4  are  used  to  check the
validity of the initial  calibration.
      Calculate the percent drift using:

                c,   -  cc
      % Drift = 	  x 100


where:
C,
      C,  =   Calibration Check Compound standard concentration.
      Cc =   Measured concentration using selected quantitation method.

      If the percent difference  for each CCC is less  than or equal to 20%,
the initial calibration is assumed to be valid.  If the criterion is not
met (>  20%  drift)  for  any one  CCC,  corrective  action must  be taken.
Problems similar to those  listed under SPCCs could affect this criterion.
If no source of the problem  can  be determined after corrective action has
been  taken,  a  new  five-point   calibration  must  be  generated.    This
criterion must  be met before sample analysis begins.   If the CCCs are not
analytes required by the permit,  then all required analytes must meet the
20% drift criterion.

      7.5.5 The  internal  standard responses and retention times  in the
calibration check standard must be evaluated immediately after or during
data acquisition.   If the retention time for any internal standard changes
by more than 30 seconds from the  last  daily calibration (Sec.  7.4), the
chromatographic system must  be inspected for malfunctions and corrections
must  be  made,  as required.    If the  EICP  area for  any  of the internal
standards changes by a factor of two  (-50% to  +100%) from the last daily
calibration check standard,  the  mass spectrometer must be  inspected for
malfunctions and corrections must be  made,  as  appropriate.

7.6   GC/MS analysis

      7.6.1 It  is  highly  recommended  that  the extract  be  screened  on a
GC/FID  or  GC/PID using the same type  of  column.   This will  minimize
contamination of the GC/MS  system from  unexpectedly high concentrations
of organic compounds.

      7.6.2 Spike the 1 mL extract obtained from sample preparation with
10 /iL of the  internal standard solution (Sec. 5.4) just  prior to analysis.

      7.6.3 Analyze the 1  ml extract by GC/MS using the  appropriate column
(as specified in Sec. 4.1.2).  The recommended GC/MS operating conditions
to be used are specified in  Sec.  7.3.
                            8250A - 12                        Revision 1
                                                          September 1994

-------
      7.6.4 If the response for any quantitation ion exceeds the initial
calibration curve range of the GC/MS  system,  extract  dilution  must take
place.  Additional  internal standard must  be added to the diluted extract
to  maintain  the required  40  ng//xL   of  each   internal  standard  in  the
extracted volume.  The diluted extract must be reanalyzed.

      7.6.5 Perform  all  qualitative  and  quantitative measurements  as
described in Sec. 7.7.  Store the extracts at 4°C,  protected from  light
in screw-cap vials  equipped with unpierced Teflon lined septa.

7,7   Data interpretation

      7.7.1 Qualitative analysis

            7.7.1.1     The   qualitative   identification   of  compounds
      determined by  this  method is  based on retention  time,  and  on
      comparison of the sample mass spectrum, after background correction,
      with  characteristic  ions   in  a  reference mass spectrum.    The
      reference mass spectrum must be generated  by  the laboratory using
      the conditions  of  this  method.   The characteristic  ions  from the
      reference mass  spectrum  are defined to be the three ions of greatest
      relative intensity, or  any  ions  over 30% relative intensity if less
      than three such  ions occur in  the reference  spectrum.   Compounds
      should be identified as present when the criteria below are met.

                  7.7.1.1.1   The intensities of the characteristic ions
            of a compound maximize in  the  same  scan or within one scan of
            each other.   Selection of  a  peak by  a data  system  target
            compound  search  routine  where  the search  is  based  on  the
            presence  of  a  target  chromatographic  peak containing  ions
            specific  for the  target  compound  at  a  compound-specific
            retention time will be accepted as meeting this criterion.

                  7.7.1.1.2   The RRT of  the  sample component  is  within
            ± 0.06 RRT units  of the RRT of the standard component.

                  7.7.1.1.3   The    relative     intensities     of    the
            characteristic  ions  agree  within  30%   of   the   relative
            intensities  of  these  ions   in   the   reference  spectrum.
            (Example:    For  an  ion  with  an   abundance  of  50%  in  the
            reference spectrum, the corresponding abundance  in  a  sample
            spectrum can range between 20% and 80%.)

                  7.7.1.1.4   Structural   isomers   that   produce   very
            similar  mass spectra  should  be   identified  as  individual
            isomers  if  they  have sufficiently  different GC  retention
            times.   Sufficient GC resolution is achieved if the height of
            the valley between two isomer peaks  is  less than 25%  of the
            sum of the two peak  heights.   Otherwise,  structural  isomers
            are identified as isomeric pairs.

                  7.7.1.1.5   Identification  is  hampered  when  sample
            components are not resolved chromatographically  and produce


                           8250A -  13                        Revision 1
                                                          September 1994

-------
      mass  spectra  containing ions  contributed  by more  than  one
      analyte.  When gas chromatographic peaks obviously represent
      more than one sample  component  (i.e.,  a broadened peak with
      shoulder(s)  or  a  valley  between  two  or  more  maxima),
      appropriate  selection  of  analyte  spectra  and  background
      spectra is  important.   Examination  of  extracted ion current
      profiles  of appropriate  ions  can aid  in  the  selection  of
      spectra, and in qualitative identification of compounds.  When
      analytes  coelute  (i.e.,  only  one  chromatographic peak  is
      apparent), the identification criteria  can  be met,  but each
      analyte spectrum will contain extraneous ions contributed by
      the coeluting compound.

      7.7.1.2     For samples containing components not associated
with the calibration standards, a library search may be made for the
purpose of tentative identification.  The necessity to perform this
type of  identification  will be determined  by the purpose  of  the
analyses  being   conducted.    Computer  generated  library  search
routines  should  not   use  normalization  routines   that   would
misrepresent the library or  unknown  spectra when  compared to each
other.   For example, the RCRA permit or waste delisting requirements
may require the  reporting of nontarget analytes.  Only after visual
comparison of sample spectra with  the nearest  library searches will
the  mass spectral  interpretation specialist  assign  a  tentative
identification.  Guidelines  for making tentative identification are:

      (1)   Relative intensities  of  major  ions  in the  reference
spectrum (ions > 10% of  the  most abundant ion) should be present in
the sample spectrum.

      (2)  The relative intensities of the major ions  should agree
within + 20%.  (Example:  For an ion with an abundance of 50% in  the
standard spectrum, the corresponding sample  ion  abundance must  be
between 30 and 70%.)

      (3)  Molecular ions  present  in  the reference spectrum should
be present in sample the spectrum.

      (4)   Ions  present  in the  sample spectrum  but  not  in  the
reference  spectrum  should  be  reviewed  for  possible  background
contamination or presence  of coeluting compounds.

      (5)   Ions  present in the reference spectrum but  not  in  the
sample spectrum  should be reviewed  for possible subtraction from the
sample spectrum  because of background  contamination  or  coeluting
peaks.  Data system  library  reduction programs can sometimes create
these discrepancies.

7.7.2 Quantitative Analysis

      7.7.2.1     When  a   compound   has  been   identified,   the
quantitation  of that  compound  will   be based  on the  integrated
abundance from the EICP of the primary characteristic ion.
                      8250A - 14                        Revision 1
                                                    September 1994

-------
       7.7.2.2      If the %RSD  of a  compound's  relative response
factor is 15% or less,  then the  concentration in  the  extract may be
determined  using the  average  response  factor  (RF)  from initial
calibration data  (Sec. 7.4.3) and the  following  equation:

                     (Ax x Cis)
       C
       ex
                     (Ais x RF)

      where Cex  is the concentration of the compound  in the extract,
and the  other terms  are as defined  in Sec.  7.4.3.

      7.7.2.3     Alternatively,  the  regression  line fitted to the
initial calibration (Sec.  7.4.6.1) may be  used for determination of
the extract concentration.

      7.7.2.4     Compute the  concentration of the  analyte in the
sample using the equations in Sees. 7.7.2.4.1 and 7.7.2.4.2.

            7.7.2.4.1   The  concentration  of  the  analyte  in the
      liquid  phase   of   the  sample   is   calculated   using  the
      concentration of the analyte in  the  extract and the volume of
      liquid extracted, as follows:
            Concentration  in liquid  (/j.g/1)  =  i_C.v x V9J
                                                   o

      where:

            Vex   =      extract  volume,  in  mL
            V0    =      volume of  liquid  extracted,  in  L.

            7.7.2.4.2    The  concentration  of the analyte  in the
      solid   phase   of  the   sample  is   calculated   using  the
      concentration of the pollutant in the extract and the weight
      of the solids, as follows:

            Concentration  in solid (M9Ag)  =  .(£.„ x VBX)
                                                   s

      where:
            Vex    =     extract  volume,  in  ml
            Ws    =     sample weight,  in kg.
      7.7.2.5     Where applicable, an estimate of concentration for
noncalibrated components in  the sample should be made.  The formulae
given above should be used  with  the following  modifications:   The
areas Ax and  Ais  should  be from the total  ion chromatograms and the
RF for the compound should  be  assumed  to  be 1.   The concentration
obtained should  be  reported indicating  (1)  that the value  is  an
estimate and  (2) which  internal  standard  was  used  to  determine
                      8250A - 15                        Revision 1
                                                    September 1994

-------
            concentration.     Use   the   nearest  internal   standard  free  of
            interferences.

                  7.7.2.6      Quantitation  of  multicomponent  compounds  (e.g.
            Aroclors)  is  beyond  the  scope   of  Method  8250A.    Normally,
            quantisation is performed using a GC/ECD by Method 8080.


8.0   QUALITY CONTROL

      8.1   Each  laboratory that  uses these methods is required  to operate a
formal quality control program.  The minimum requirements of this program consist
of an initial  demonstration of laboratory  capability and an ongoing analysis of
spiked  samples  to evaluate and  document data  quality.   The  laboratory  must
maintain records to document  the quality  of the data generated.   Ongoing  data
quality checks are compared with established performance criteria to determine
if the results of analyses meet the performance characteristics of the method.
When results of sample spikes  indicate  atypical  method performance, a quality
control check standard must be  analyzed  to  confirm  that  the  measurements  were
performed in an  in-control  mode of operation.

      8.2   Before  processing  any  samples,  the  analyst   should  demonstrate,
through  the  analysis  of a reagent  water blank,  that  interferences  from  the
analytical  system, glassware,  and reagents are under control.  Each time a set
of samples  is  extracted  or  there is  a  change in  reagents, a reagent water blank
should be processed as a  safeguard against chronic laboratory contamination.  The
blank samples  should be carried through all stages of the sample preparation and
measurement steps.

      8.3   The  experience  of  the  analyst  performing  GC/MS  analyses  is
invaluable  to  the  success of the methods.  Each  day that analysis is performed,
the  daily   calibration   standard  should  be  evaluated to   determine  if  the
chromatographic  system is  operating properly.   Questions  that  should be asked
are:   Do the  peaks look normal?;  Is  the response obtained  comparable  to  the
response from previous  calibrations?   Careful examination  of the  standard
chromatogram can  indicate  whether the column  is still good, the  injector is
leaking, the injector septum needs replacing,  etc.   If  any changes are made to
the system (e.g. column  changed),  recal ibration of the  system must take place.

      8.4   Required instrument QC is found in  the following section:

            8.4.1 The   GC/MS   system  must  be  tuned  to  meet   the   DFTPP
      specifications in  Sec. 7.3.1 and 7.4.1.

            8.4.2 There must  be an  initial  calibration of  the  GC/MS system as
      specified in Sec.  7.4.

            8.4.3 The GC/MS system must meet the SPCC criteria specified in Sec.
      7.5.3 and the CCC  criteria in Sec. 7.5.4, each 12 hr.

      8.5   To  establish  the  ability  to  generate  acceptable  accuracy  and
precision,  the analyst must perform the following operations.
                                  8250A - 16                        Revision 1
                                                                September 1994

-------
            8.5.1 A  quality  control  (QC)  check sample concentrate is required
      containing each analyte  at  a  concentration  of 100 mg/L in acetone.  The
      QC check sample concentrate may be prepared from pure standard materials
      or purchased as certified solutions.  If prepared by the laboratory, the
      QC check sample concentrate must  be  made using stock standards prepared
      independently  from those used for calibration.

            8.5.2 Using a pipet, prepare QC check  samples  at a concentration of
      100 jug/L by adding 1.00 ml of QC check sample  concentrate to each of four
      1-L aliquots of organic-free reagent water.

            8.5.3 Analyze  the well-mixed  QC  check  samples  according  to  the
      method beginning in Sec. 7.1 with extraction of the samples.

            8.5.4 Calculate  the average recovery (x) in jiig/L, and the standard
      deviation of the  recovery  (s)  in /Ltg/L, for each  analyte  using the four
      results.

            8.5.5 For  each  analyte  compare  s  and  x  with the  corresponding
      acceptance criteria jfor precision and  accuracy,  respectively,  found in
      Table 6.   If   s and x  for  all  analytes of interest  meet  the acceptance
      criteria, the  system performance is  acceptable and analysis  of actual
      samples can begin.   If  any individual s exceeds the precision limit or any
      individual   x   falls  outside the  range  for  accuracy,  then the  system
      performance is unacceptable for that analyte.

            NOTE: The large number of analytes in Table  6 present a substantial
                  probability  that  one  or more will fail  at  least  one of the
                  acceptance  criteria when  all  analytes of a given method are
                  analyzed.

            8.5.6 When one or more of the analytes tested fail at least one of
      the  acceptance criteria, the  analyst  must proceed according  to  Sees.
      8.5.6.1 or 8.5.6.2.

                  8.5.6.1      Locate  and correct the source of the problem and
            repeat the  test  for all  analytes of interest  beginning  with Sec.
            8.5.2.

                  8.5.6.2      Beginning with  Sec. 8.5.2,  repeat  the  test only
            for those analytes that  failed  to  meet criteria.  Repeated failure,
            however, will confirm  a general problem with  the measurement system.
            If this  occurs,  locate  and correct the  source of  the problem and
            repeat the test  for all  compounds of  interest beginning  with Sec.
            8.5.2.

      8.6   The laboratory must,  on  an  ongoing  basis,  analyze a method blank,
a matrix spike, and  a matrix  spike/duplicate for each analytical batch (up to a
maximum of 20  samples/batch)  to assess accuracy.  For  laboratories analyzing one
to ten samples per month, at least one spiked sample per month is required.
                                  8250A - 17                        Revision 1
                                                                September 1994

-------
      8.6.1 The  concentration  of. the  spike  in  the  sample  should  be
determined as follows:

            8.6.1.1      If, as in compliance monitoring, the concentration
      of  a specific  analyte  in the  sample  is being  checked  against a
      regulatory concentration  limit,  the  spike should be at that limit
      or 1 to 5 times higher than the  background concentration determined
      in Sec. 8.6.2, whichever  concentration would be  larger.

            8.6.1.2      If the  concentration of a  specific analyte in the
      sample  is not  being  checked   against  a limit  specific  to  that
      analyte, the spike should  be  at  100 /ng/L or 1 to 5 times higher than
      the  background  concentration determined  in  Sec.  8.6.2,  whichever
      concentration would be larger.

            8.6.1.3      If  it  is  impractical  to determine  background
      levels  before  spiking   (e.g.,  maximum  holding  times  will   be
      exceeded), the spike concentration should be at (1) the regulatory
      concentration limit,  if any; or,  if  none (2)  the larger of either
      5  times  higher  than  the  expected   background   concentration  or
      100
      8.6.2 Analyze  one  sample  aliquot  to  determine  the  background
concentration (B) of each analyte.  If necessary, prepare a new QC check
sample   concentrate   (Sec.   8.5.1)   appropriate  for   the   background
concentration in the sample.  Spike a second sample aliquot with 1.00 ml
of the QC  reference  sample  concentrate and analyze  it  to  determine the
concentration after spiking (A) of each analyte.  Calculate each percent
recovery (p) as  100(A-B)%/T, where T is the known true value of the spike.

      8.6.3 Compare the  percent  recovery (p)  for each  analyte  with the
corresponding QC acceptance criteria found in  Table 6.  These acceptance
criteria were calculated to  include  an  allowance  for error in measurement
of both  the background  and spike  concentrations,  assuming  a  spike  to
background ratio of 5:1.  This error will be accounted for to the extent
that the analyst's spike to background ratio approaches 5:1.   If spiking
was performed at a concentration lower than 100 /xg/L,  the analyst must use
either the  QC acceptance criteria presented in  Table 6, or  optional  QC
acceptance criteria calculated for the specific spike concentration.  To
calculate optional  acceptance criteria  for the  recovery of an analyte: (1)
Calculate accuracy (x')  using  the  equation  found  in Table 7, substituting
the spike concentration  (T) for C;  (2)  calculate overall  precision (S')
using the equation in Table 7,  substituting x'  for  x; (3)  calculate the
range   for   recovery    at   the   spike   concentration   as   (100x'/T)
± 2.44(100S'/T)%.

      8.6.4  If  any  individual  p falls outside  the  designated  range for
recovery,  that  analyte  has  failed the  acceptance  criteria.   A  check
standard containing each  analyte that failed the  criteria must be analyzed
as described in  Sec.  8.7.
                            8250A -  18                        Revision 1
                                                          September 1994

-------
      8.7    If  any  analyte  fails  the acceptance criteria for recovery in Sec.
8.6, a QC check  standard containing each analyte  that failed must be prepared and
analyzed.

      NOTE:  The frequency for the required analysis of a  QC check  standard will
             depend upon the number of analytes being simultaneously tested, the
             complexity  of  the  sample matrix,  and  the  performance of  the
             laboratory.   If the  entire list of  analytes  in Table  6 must be
             measured  in  the  sample  in Sec.  8.6, the  probability  that  the
             analysis of a QC check standard will be required is high.  In this
             case, the QC  check  standard should  be routinely  analyzed with the
             spiked sample.

             8.7.1 Prepare the  QC  reference sample by  adding 1.0  ml  of the QC
      check  sample concentrate  (Sec.  8.5.1  or 8.6.2)  to  1  L  of reagent water.
      The QC check standard needs  only to  contain  the analytes  that failed
      criteria  in the test in Sec. 8.6.

             8.7.2 Analyze the QC check  standard to determine the concentration
      measured  (A)  of  each  analyte.    Calculate  each  percent recovery (PJ  as
      100(A/T)%, where T is the true value of the standard concentration.

             8.7.3 Compare the  percent recovery  (PJ for  each analyte with the
      corresponding QC acceptance  criteria  found in Table 6.   Only  analytes that
      failed the test in Sec. 8.6 need to be compared with these criteria.  If
      the recovery of any such analyte falls outside the designated range, the
      laboratory performance for that  analyte is  judged  to  be  out of control,
      and the problem must be immediately  identified and  corrected.  The result
      for that analyte in  the unspiked sample is  suspect and may  not be reported
      for regulatory compliance purposes.

      8.8    As  part of the  QC  program for the  laboratory,  method accuracy for
each matrix studied must  be  assessed  and records must be  maintained.  After the
analysis of  five spiked samples _(of the same matrix) as  in Sec. 8.6,  calculate
the average  percent recovery  (p)  and the  standard  deviation   of  the percent
recovery (sp).   Express  the  accuracy  assessment  as a percent  recovery interval
from p  - 2sp to p +  2sp.   If p = 90% and  sp =  10%, for  example,  the accuracy
interval is  expressed  as 70-110%.   Update  the accuracy assessment  for each
analyte  on   a   regular  basis   (e.g.   after  each  five  to   ten  new  accuracy
measurements).

      8.9    To  determine acceptable accuracy and precision limits for surrogate
standards the following procedure should be performed.

             8.9.1 For each  sample  analyzed,  calculate  the  percent recovery of
      each surrogate in the sample.

             8.9.2 Once a minimum of thirty samples of the same matrix have been
      analyzed,  calculate  the  average  percent   recovery   (P)  and  standard
      deviation of the percent recovery (s) for each of the surrogates.
                                  8250A - 19                        Revision 1
                                                                September 1994

-------
            8.9.3 For  a  given matrix,  calculate  the upper and  lower control
      limit for method performance for each  surrogate standard.  This should be
      done as follows:

            Upper Control Limit (UCL) = P + 3s
            Lower Control Limit (LCL) = P - 3s

            8.9.4 For  aqueous  and soil  matrices,  these laboratory established
      surrogate  control  limits should,  if applicable,  be compared  with  the
      control limits listed in Table 8.  The limits given in Table 8 are multi-
      laboratory performance  based limits for  soil  and aqueous  samples,  and
      therefore, the  single-laboratory  limits established in  Step  8.9.3 must
      fall within those given in Table 8 for these matrices.

            8.9.5 If recovery is not within limits, the following procedures are
      required.

            •     Check  to  be sure  there  are  no   errors  in  calculations,
                  surrogate  solutions  and  internal   standards.   Also,  check
                  instrument performance.

            •     Recalculate  the  data  and/or reanalyze  the  extract if any of
                  the  above checks  reveal a problem.

            •     Reextract and reanalyze the sample if  none  of the above are
                  a problem or  flag the data  as "estimated concentration".

            8.9.6 At a minimum, each laboratory should update surrogate recovery
      limits on a matrix-by-matrix basis,  annually.

      8.10  It  is  recommended  that the  laboratory  adopt additional  quality
assurance practices for use with this method.   The specific practices that are
most productive depend upon the needs of  the  laboratory  and  the nature of the
samples.   Field duplicates  may be  analyzed  to  assess  the  precision  of  the
environmental measurements.  When  doubt exists over the identification of a peak
on the chromatogram, confirmatory techniques such as gas chromatography with a
dissimilar column or mass spectrometry using other  ionization modes  must be used.
Whenever possible, the laboratory should  analyze standard reference materials and
participate in relevant performance evaluation studies.


9.0   METHOD PERFORMANCE

      9.1   Method 8250 was tested by 15 laboratories  using organic-free reagent
water, drinking water, surface water, and industrial  wastewaters spiked at six
concentrations  over  the  range  5-1,300 ^g/L.   Single  operator  accuracy  and
precision,  and method  accuracy  were  found  to  be  directly  related  to  the
concentration of the analyte and essentially independent of the sample matrix.
Linear equations to describe these relationships are presented in Table 7.
                                  8250A - 20                        Revision 1
                                                                September 1994

-------
10.0  REFERENCES

1.    U.S. EPA 40 CFR  Part 136, "Guidelines Establishing Test  Procedures for the
      Analysis of Pollutants Under the Clean Water Act,  Method 625," October 26,
      1984.

2.    U.S.  EPA Contract  Laboratory Program,  Statement of  Work  for  Organic
      Analysis, July  1985, Revision.

3.    Eichelberger, J.W., L.E. Harris,  and  W.L.  Budde,  "Reference Compound to
      Calibrate   Ion   Abundance   Measurement   in  Gas   Chromatography-Mass
      Spectrometry Systems," Analytical Chemistry, 47,  995-1000, 1975.

4.    "Method Detection Limit for Methods  624  and  625," Olynyk,  P., W.L. Budde,
      and J.W. Eichelberger, Unpublished report, October 1980.

5.    "Interlaboratory Method Study  for EPA Method  625-Base/Neutrals, Acids, and
      Pesticides," Final Report  for EPA Contract  68-03-3102  (in preparation).

6.    Burke,  J.A.  "Gas  Chromatography  for  Pesticide  Residue  Analysis;  Some
      Practical  Aspects,"  Journal   of  the  Association  of  Official  Analytical
      Chemists, 48, 1037, 1965.
                                  8250A - 21                        Revision 1
                                                                September 1994

-------
                        TABLE 1.
CHROMATOGRAPHIC CONDITIONS, METHOD DETECTION LIMITS,  AND
     CHARACTERISTIC IONS FOR SEMIVOLATILE COMPOUNDS
Compound
Acenaphthene
Acenaphthene-d10 (I.S.)
Acenaphthylene
Acetophenone
Aldrin
4-Aminobiphenyl
Aniline
Anthracene
Aroclor-1016b
Aroclor-1221b
Aroclor-1232b
Aroclor-1242b
Aroclor-1248b
Aroclor-1254b
Aroclor-1260b
Benzidine"
Benzoic acid
Benzo(a) anthracene
Benzo(b)fl uoranthene
Benzo(k)fl uoranthene
Benzo(g,h,i)perylene
Benzo(a)pyrene
Benzyl alcohol
a-BHCa
iS-BHC
5-BHC
7-BHC (Lindane)9
Bis(2-chloroethoxy)methane
Bis(2-chloroethyl) ether
Bis(2-chloroisopropyl ) ether
Bis(2-ethylhexyl) phthalate
4-Bromophenyl phenyl ether
Butyl benzyl phthalate
Chlordane6
4-Chloroaniline
1-Chloronaphthalene
2-Chloronaphthalene
4-Chl oro-3-methyl phenol
2-Chlorophenol
4-Chlorophenyl phenyl ether
Chrysene
Chrysene-d12 (I.S.)
4,4'-DDD
Method
Retention Detection Primary Secondary
Time (min) limit (/zg/L) Ion Ion(s)
17.8
--
17.4
--
24.0
--
--
22.8
18-30
15-30
15-32
15-32
12-34
22-34
23-32
28.8
--
31.5
34.9
34.9
45.1
36.4
--
21.1
23.4
23.7
22.4
12.2
8.4
9.3
30.6
21.2
29.9
19-30
--
_.
15.9
13.2
5.9
19.5
31.5
--
28.6
1.9
--
3.5
--
1.9
--
--
1.9

30



36
--
44
--
7.8
4.8
2.5
4.1
2.5
--
--
4.2
3.1
--
5.3
5.7
5.7
2.5
1.9
2.5
--
--
--
1.9
3.0
3.3
4.2
2.5
--
2.8
154
164
152
105
66
169
93
178
222
190
190
222
292
292
360
184
122
228
252
252
276
252
108
183
181
183
183
93
93
45
149
248
149
373
127
162
162
107
128
204
228
240
235
153, 152
162, 160
151, 153
77, 51
263, 220
168, 170
66, 65
176, 179
260, 292
224, 260
224, 260
256, 292
362, 326
362, 326
362, 394
92, 185
105, 77
229, 226
253, 125
253, 125
138, 277
253, 125
79, 77
181, 109
183, 109
181, 109
181, 109
95, 123
63, 95
77, 121
167, 279
250, 141
91, 206
375, 377
129
127, 164
127, 164
144, 142
64, 130
206, 141
226, 229
120, 236
237, 165
                       8250A -  22
    Revision 1
September 1994

-------
TABLE 1.
(Continued)
Compound
4,4'-DDT
4,4'-DDE
Dibenz(a,j)acridine
Dibenz( a, h) anthracene
Dibenzofuran
Di-n-butyl phthalate
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
l,4-Dichlorobenzene-d4 (I.S.)
3,3'-Dichlorobenzidine
2,4-Dichlorophenol
2,6-Dichlorophenol
Dieldrin
Diethyl phthalate
p-Dimethyl aminoazobenzene
7, 12-Dimethylbenz(a)anthracene
a-,a-Dimethylphenethylamine
2,4-Dimethylphenol
Dimethyl phthalate
4, 6-Dinitro- 2 -methyl phenol
2,4-Dinitrophenol
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Diphenylamine
1,2-Diphenylhydrazine
Di-n-octyl phthalate
Endosulfan Ia
Endosulfan IT
Endosulfan sulfate
Endrina
Endrin aldehyde
Endrin ketone
Ethyl methanesulfonate
Fluoranthene
Fluorene
2-Fluorobiphenyl (surr.)
2-Fluorophenol (surr.)
Heptachlor
Heptachlor epoxide
Hexachlorobenzene
Hexachlorobutadiene
Hexachl orocycl opentadi ene"
Hexachloroethane
Method
Retention Detection Primary Secondary
Time (min) Limit (/zg/L) Ion Ion(s)
29.3
27.2

43.2

24.7
8.4
7.4
7.8
--
32.2
9.8
--
27.2
20.1
--
--
--
9.4
18.3
16.2
15.9
19.8
18.7
--
--
32.5
26.4
28.6
29.8
27.9
--
--
--
26.5
19.5
--
--
23.4
25.6
21.0
11.4
13.9
8.4
4.7
--
--
2.5
--
2.5
1.9
1.9
4.4
--
16.5
2.7
--
2.5
1.9
--
--
--
2.7
1.6
24
42
5.7
1.9

--
2.5
--
--
5.6
--
--
--
--
2.2
1.9
--
--
1.9
2.2
1.9
0.9
--
1.6
235
246
279
278
168
149
146
146
146
152
252
162
162
79
149
120
256
58
122
163
198
184
165
165
169
77
149
195
337
272
263
67
317
79
202
166
172
112
100
353
284
225
237
117
237,
24,
280,
139,
139
150,
148,
148,
148,
150,
254,
164,
164,
263,
177,
225,
241,
91,
107,
194,
51,
63,
63,
63,
168,
105,
167,
339,
339,
387,
82,
345,
67,
109,
101,
165,
171
64
272,
355,
142,
223,
235,
201,
165
176
277
279

104
111
111
111
115
126
98
98
279
150
77
257
42
121
164
105
154
89
89
167
182
43
341
341
422
81
250
319
97
203
167


274
351
249
227
272
199
8250A - 23
    Revision 1
September 1994

-------
TABLE 1.
(Continued)
Compound
Indeno(l,2,3-cd)pyrene
Isophorone
Methoxychlor
3-Methyl chol anthrene
Methyl methanesulfonate
2 -Methyl naphthalene
2-Methyl phenol
4-Methyl phenol
Naphthalene
Naphthalene-d8 (I.S.)
1-Naphthylamine
2-Naphthylamine
2-Nitroaniline
3-Nitroanil ine
4-Nitroanil ine
Nitrobenzene
Nitrobenzene-d5 (surr.)
2-Nitrophenol
4-Nitrophenol
N-Nitroso-di -n-butylamine
N-Nitrosodi methyl ami nea
N-Nitrosodiphenylaminea
N-Nitroso-di-n-propylamine
N-Nitrosopi peri dine
Pentachl orobenzene
Pentachloronitrobenzene
Pentachlorophenol
Perylene-d12 (I.S.)
Phenacetin
Phenanthrene
Phenanthrene-d10 (I.S.)
Phenol
Phenol -de (surr.)
2-Picoline
Pronamide
Pyrene
Terphenyl-d14 (surr.)
1,2,4, 5-Tetrachl orobenzene
2,3,4,6-Tetrachlorophenol
Method
Retention Detection Primary Secondary
Time (min) Limit (p.g/1) Ion Ion(s)
42.7
11.9
--
__
--
--
--
--
12.1
--
--
--
--
--
--
11.1
--
6.5
20.3
--
--
20.5
--
--
--
--
17.5
--
--
22.8
--
8.0
--
--
--
27.3
--
--
--
3.7
2.2
--
--
--
--
--
--
1.6
--
--
--
--
--
--
1.9
--
3.6
2.4
--
--
1.9
--
--
--
--
3.6
--
--
5.4
--
1.5
--
--

1.9
--
--
--
276
82
227
268
80
142
108
108
128
136
143
143
65
138
138
77
82
139
139
84
42
169
70
42
250
295
266
264
108
178
188
94
99
93
173
202
244
216
232
138,
95,
228
253,
79,
141
107,
107,
129,
68
115,
115,
92,
108,
108,
123,
128,
109,
109,
57,
74,
168,
130,
114,
252,
237,
264,
260,
109,
179,
94,
65,
42,
66,
175,
200,
122,
214,
230,
227
138

267
65

79
79
127

116
116
138
92
92
65
54
65
65
41
44
167
42
55
248
142
268
265
179
176
80
66
71
92
145
203
212
218
131
8250A - 24
    Revision 1
September 1994

-------
TABLE 1.
(Continued)


Compound
Toxapheneb
2,4,6-Tribromophenol (surr.)
1,2,4-Trichlorobenzene
2,4,5-Trichlorophenol
2,4,6-Trichlorophenol

Retention
Time (min)
25-34
--
11.6
--
11.8
Method
Detection
Limit (MQ/L)
— —
--
1.9
--
2.7

Primary
Ion
159
330
180
196
196

Secondary
Ion(s)
231, 233
332, 141
182, 145
198, 200
198, 200
"See  Sec.  1.3
bThese compounds are mixtures of various isomers.
(I.S.)  = Internal Standard
(surr). = Surrogate
                                   TABLE 2.
             DETERMINATION OF ESTIMATED QUANTITATION LIMITS (EQL)
                             FOR VARIOUS MATRICES8
Matrix
  Factor
Ground water                                                           10
Low-concentration soil by ultrasonic extraction with GPC cleanup      670
High-concentration soil and sludges by ultrasonic extraction       10,000
Non-water miscible waste                                          100,000
    EQL  =  [Method detection limit (see Table 1)]  X  [Factor  found  in this table].
    For  non-aqueous samples,  the factor is on a wet-weight basis.  Sample EQLs
    are  highly  matrix-dependent.  The EQLs  to be determined herein are provided
    for  guidance  and may  not  always  be  achievable.
                                  8250A - 25
    Revision 1
September 1994

-------
                                   TABLE 3.
                  DFTPP KEY IONS AND ION ABUNDANCE CRITERIA"
       Mass             Ion Abundance Criteria
       51               30-60% of mass 198

       68               < 2% of mass 69
       70               < 2% of mass 69

      127               40-60% of mass 198

      197               < 1% of mass 198
      198               Base peak, 100% relative abundance
      199               5-9% of mass 198

      275               10-30% of mass 198

      365               > 1% of mass 198

      441               Present but less than mass 443
      442               > 40% of mass 198
      443               17-23% of mass 442
"See  Reference 3.
                                  8250A - 26                        Revision  1
                                                                September  1994

-------
                             TABLE 4.
                    CALIBRATION  CHECK COMPOUNDS
Base/Neutral Fraction                   Acid Fraction
Acenaphthene                            4-Chloro-3-methylphenol
1,4-Dichlorobenzene                     2,4-Dichlorophenol
Hexachlorobutadiene                     2-Nitrophenol
N-Nitroso-di-n-phenylamine              Phenol
Di-n-octyl phthalate                    Pentachlorophenol
Benzo(a)pyrene                          2,4,6-Trichlorophenol
Fluoranthene
                            8250A - 27                        Revision 1
                                                          September 1994

-------
                                   TABLE 5.
          SEMIVOLATILE  INTERNAL  STANDARDS  WITH CORRESPONDING  ANALYTES
                           ASSIGNED  FOR  QUANTITATION
1,4-Dichlorobenzene-D,,
Naphthalene-d8
Acenaphthene-d10
Aniline
Benzyl alcohol
Bis(2-chloroethyl) ether
Bis(2-chloroisopropyl)ether
2-Chlorophenol
1,3-Dichlorobenzene
1,4-Dichlorobenzene
1,2-Di chlorobenzene
Ethyl methanesulfonate
2-Fluorophenol (surr.)
Hexachloroethane
Methyl methanesulfonate
2-Methylphenol
4-Methylphenol
N-Nitrosodimethylamine
N-Nitroso-di-n-propylamine
Phenol
Phenol-d6 (surr.)
2-Picoline
Acetophenone
Benzole acid
Bis(2-chloroethoxy)methane
4-Chloroaniline
4-Chloro-3-methyl phenol
2,4-Dichlorophenol
2,6-Dichlorophenol
a,a-Dimethylphenethylamine
2,4-Dimethylphenol
Hexachlorobutadiene
Isophorone
2-Methylnaphthalene
Naphthalene
Nitrobenzene
Nitrobenzene-d8 (surr.)
2-Nitrophenol
N-Nitroso-di-n-butyl amine
N-Nitrosopiperidine
1,2,4-Trichlorobenzene
Acenaphthene
Acenaphthylene
1-Chloronaphthalene
2-Chloronaphthalene
4-Chlorophenyl
  phenyl  ether
Dibenzofuran
Diethyl phthalate
Dimethyl  phthalate
2,4-Dinitrophenol
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Fluorene
2-Fluorobiphenyl
  (surr.)
Hexachlorocyclo-
  pentadiene
1-Naphthylamine
2-Naphthylamine
2-Nitroaniline
3-Nitroani1ine
4-Nitroaniline
4-Nitrophenol
Pentachlorobenzene
1,2,4,5-Tetrachloro-
  benzene
2,3,4,6-Tetrachloro-
  phenol
2,4,6-Tribromophenol
  (Surr.)
2,4,6-Trichlorophenol
2,4,5-Trichlorophenol
(surr.) = surrogate
                                  8250A - 28
                                        Revision 1
                                    September 1994

-------
                                   TABLE 5.
          SEMIVOLATILE INTERNAL STANDARDS WITH CORRESPONDING ANALYTES
                           ASSIGNED FOR QUANTITATION
                                  (Continued)
Phenanthrene-d
              10
Chrysene-d12
Perylene-d
                                                                   12
4-Aminobiphenyl
Anthracene
4-Bromophenyl phenyl ether
Di-n-butyl phthalate
4,6-Dinitro-2-methyl phenol
Diphenylamine
1,2-Di phenylhydrazi ne
Fluoranthene
Hexachlorobenzene
N-Nitrosodiphenylamine
Pentachlorophenol
Pentachloronitrobenzene
Phenacetin
Phenanthrene
Pronamide
Benzidine
Benzo(a)anthracene
Bis(2-ethylhexyl) phthalate
Butyl benzyl phthalate
Chrysene
3,3'-Dichlorobenzidine
p-Dimethylaminoazobenzene
Pyrene
Terphenyl-d14 (surr.)
Benzo(b)fluoranthene
Benzo(k) fl uoranthene
Benzo(g,h,i)perylene
Benzo(a)pyrene
Dibenz(a,j)acridine
Dibenz(a,h)anthracene
7,12-Dimethylbenz-
  (a)anthracene
Di-n-octyl phthalate
Indeno(l,2,3-cd)pyrene
3-Methylcholanthrene
(surr,) = surrogate
                                  8250A - 29
                                        Revision 1
                                    September 1994

-------
       TABLE 6.
QC ACCEPTANCE CRITERIA"
Compound
Acenaphthene
Acenaphthylene
Aldrin
Anthracene
Benzo(a)anthracene
Benzo(b)fluoranthene
Benzo(k)fluoranthene
Benzo(a)pyrene
Benzo(ghi)perylene
Butyl benzyl phthalate
6-BHC
(5-BHC
Bis(2-chloroethy1) ether
Bis(2-chloroethoxy)methane
Bis(2-chloroisopropyl) ether
Bis(2-ethylhexyl) phthalate
4-Bromophenyl phenyl ether
2-Chloronaphthalene
4-Chlorophenyl phenyl ether
Chrysene
4,4'-DDD
4,4'-DDE
4,4'-DDT
Dibenzo (a, h) anthracene
Di-n-butyl phthalate
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
3,3'-Dichlorobenzidine
Dieldrin
Diethyl phthalate
Dimethyl phthalate
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Di-n-octyl phthalate
Endosulfan sulfate
Endrin aldehyde
Fluoranthene
Fluorene
Heptachlor
Heptachlor epoxide
Hexachlorobenzene
Hexachl orobutadi ene
Test
cone.
(M9/L)
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
Limit
for s
(M9/L)
27.6
40.2
39.0
32.0
27.6
38.8
32.3
39.0
58.9
23.4
31.5
21.6
55.0
34.5
46.3
41.1
23.0
13.0
33.4
48.3
31.0
32.0
61.6
70.0
16.7
30.9
41.7
32.1
71.4
30.7
26.5
23.2
21.8
29.6'
31.4
16.7
32.5
32.8
20.7
37.2
54.7
24.9
26.3
Range
for x
(M9/L)
60.1-132.3
53.5-126.0
7.2-152.2
43.4-118.0
41.8-133.0
42.0-140.4
25.2-145.7
31.7-148.0
D-195.0
D-139.9
41.5-130.6
D-100.0
42.9-126.0
49.2-164.7
62.8-138.6
28.9-136.8
64.9-114.4
64.5-113.5
38.4-144.7
44.1-139.9
D-134.5
19.2-119.7
D-170.6
D-199.7
8.4-111.0
48.6-112.0
16.7-153.9
37.3-105.7
8.2-212.5
44.3-119.3
D-100.0
D-100.0
47.5-126.9
68.1-136.7
18.6-131.8
D-103.5
D-188.8
42.9-121.3
71.6-108.4
D-172.2
70.9-109.4
7.8-141.5
37.8-102.2
Range
P. Ps
(%)
47-145
33-145
D-166
27-133
33-143
24-159
11-162
17-163
D-219
D-152
24-149
D-110
12-158
33-184
36-166
8-158
53-127
60-118
25-158
17-168
D-145
4-136
D-203
D-227
1-118
32-129
D-172
20-124
D-262
29-136
D-114
D-112
39-139
50-158
4-146
D-107
D-209
26-137
59-121
D-192
26-155
D-152
24-116
      8250A  - 30
    Revision 1
September 1994

-------
                             TABLE 6.
                      QC  ACCEPTANCE CRITERIA"
                            (Continued)


Compound
Hexachloroethane
Indeno(l,2,3-cd)pyrene
Isophorone
Naphthalene
Nitrobenzene
N-Nitroso-di-n-propylamine
PCB-1260
Phenanthrene
Pyrene
1,2,4-Trichlorobenzene
4-Chloro-3 -methyl phenol
2-Chlorophenol
2,4-Chlorophenol
2,4-Dimethylphenol
2,4-Dinitrophenol
2-Methyl -4,6-dinitrophenol
2-Nitrophenol
4-Nitrophenol
Pentachlorophenol
Phenol
2,4,6-Trichlorophenol
Test
cone.
(M9/L)
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
Limit
for s
(^g/L)
24.5
44.6
63.3
30.1
39.3
55.4
54.2
20.6
25.2
28.1
37.2
28.7
26.4
26.1
49.8
93.2
35.2
47.2
48.9
22.6
31.7
s = Standard deviation of four recovery
x = Average recovery
p, ps = Percent recovery
D = Detected; result
a r v> -i 4- n\* •! •> ft»nm Af\ TCD D
for four
measured
must be
•,.*+ I'SC 4
recovery
•
Range
for x
(M9/L)
55.2-100.0
D-150.9
46.6-180.2
35.6-119.6
54.3-157.6
13.6-197.9
19.3-121.0
65.2-108.7
69.6-100.0
57.3-129.2
40.8-127.9
36.2-120.4
52.5-121.7
41.8-109.0
D-172.9
53.0-100.0
45.0-166.7
13.0-106.5
38.1-151.8
16.6-100.0
52.4-129.2
measurements,
measurements,

Range
P> Ps
(%)
40-113
D-171
21-196
21-133
35-180
D-230
D-164
54-120
52-115
44-142
22-147
23-134
39-135
32-119
D-191
D-181
29-182
D-132
14-176
5-112
37-144
in M9/L.
in jug/L-

greater than zero.
"f\i/« M *-i 4- U /-\ /-I
ao c TU<-» r- f* r*\
^-i+-r»v*-i^ -sv»n K-acarl
directly on the method  performance data  in Table 7.  Where necessary,  the
limits for  recovery  have  been  broadened to assure  applicability  of  the
limits to concentrations below those used to develop Table 7.
                            8250A - 31
    Revision 1
September 1994

-------
                          TABLE 7.
METHOD ACCURACY AND PRECISION AS FUNCTIONS OF CONCENTRATION8
Parameter
Acenaphthene
Acenaphthylene
Aldrin
Anthracene
Benzo( a) anthracene
Chloroethane
Benzo ( b) fl uoranthene
Benzo(k)fluoranthene
Benzo(a)pyrene
Benzo(ghi)perylene
Butyl benzyl phthalate
B-BHC
£-BHC
Bis(2-chloroethyl) ether
Bis(2-chloroethoxy) methane
Bis(2-chloroisopropyl) ether
Bis(2-ethylhexyl) phthalate
4-Bromophenyl phenyl ether
2-Chloronaphthalene
4-Chlorophenyl phenyl ether
Chrysene
4,4'-DDD
4,4'-DDE
4,4'-DDT
Dibenzo (a, h) anthracene
Di-n-butyl phthalate
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
3,3'-Dichlorobenzidine
Dieldrin
Diethyl phthalate
Dimethyl phthalate
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Di-n-octyl phthalate
Endosulfan sulfate
Endrin aldehyde
Fluoranthene
Fluorene
Heptachlor
Heptachlor epoxide
Hexachlorobenzene
Hexachlorobutadiene
Hexachloroethane
Accuracy, as
recovery, x'
(M9/L)
0.96C+0.19
0.89C+0.74
0.78C+1.66
0.80C+0.68
0.88C-0.60
0.99C-1.53
0.93C-1.80
0.87C-1.56
0.90C-0.13
0.98C-0.86
0.66C-1.68
0.87C-0.94
0.29C-1.09
0.86C-1.54
1.12C-5.04
1.03C-2.31
0.84C-1.18
0.91C-1.34
0.89C+0.01
0.91C+0.53
0.93C-1.00
0.56C-0.40
0.70C-0.54
0.79C-3.28
0.88C+4.72
0.59C+0.71
0.80C+0.28
0.86C-0.70
0.73C-1.47
1.23C-12.65
0.82C-0.16
0.43C+1.00
0.20C+1.03
0.92C-4.81
1.06C-3.60
0.76C-0.79
0.39C+0.41
0.76C-3.86
0.81C+1.10
0.90C-0.00
0.87C-2.97
0.92C-1.87
0.74C+0.66
0.71C-1.01
0.73C-0.83
Single analyst
precision, s/
(M9/L)
0.15X-0.12
0.24X-1.06
0.27X-1.28
0.21X-0.32
0.15X+0.93
0.14X-0.13
0.22X+0.43
0.19X+1.03
0.22X+0.48
0.29X+2.40
O.lSx+0.94
0.20X-0.58
0.34X+0.86
0.35X-0.99
0.16X+1.34
0.24X+0.28
0.26X+0.73
0.13X+0.66
0.07X+0.52
0.20X-0.94
0.28X+0.13
0.29X-0.32
0.26X-1.17
0.42X+0.19
0.30X+8.51
O.lSx+1.16
0.20X+0.47
0.25X+0.68
0.24X+0.23
0.28X+7.33
0.20X-0.16
0.28X+1.44
0.54x+0'.19
0.12X+1.06
0.14X+1.26
0.21X+1.19
0.12X+2.47
O.lSx+3.91
0.22X-0.73
0.12X+0.26
0.24X-0.56
0.33X-0.46
0.18X-0.10
0.19X+0.92
0.17X+0.67
Overall
precision,
S' (M9A)
0.21X-0.67
0.26X-0.54
0.43X+1.13
0.27X-0.64
0.26X-0.21
0.17X-0.28
0.29X+0.96
0.35X+0.40
0.32X+1.35
0.51X-0.44
0.53X+0.92
0.30X+1.94
0.93X-0.17
0.35X+0.10
0.26X+2.01
0.25X+1.04
0.36X+0.67
0.16X+0.66
0.13X+0.34
0.30X-0.46
0.33X-0.09
0.66X-0.96
0.39X-1.04
0.65X-0.58
0.59X+0.25
0.39X+0.60
0.24X+0.39
0.41X+0.11
0.29X+0.36
0.47X+3.45
0.26X-0.07
0.52X+0.22
1.05X-0.92
0.21X+1.50
0.19X+0.35
0.37X+1.19
0.63X-1.03
0.73X-0.62
0.28X-0.60
0.13X+0.61
O.SOx-0.23
0.28X+0.64
0.43X-0.52
0.26X+0.49
0.17X+0.80
                         8250A - 32
    Revision 1
September 1994

-------
                                   TABLE 7.
         METHOD  ACCURACY  AND  PRECISION  AS  FUNCTIONS  OF CONCENTRATION8
                                  (Continued)
Parameter
Indeno(l,2,3-cd)pyrene
Isophorone
Naphthalene
Nitrobenzene
N-Nitroso-di-n-propylamine
PCB-1260
Phenanthrene
Pyrene
1,2,4-Trichlorobenzene
4-Chloro-3 -methyl phenol
2-Chlorophenol
2,4-Dichlorophenol
2,4-Dimethylphenol
2,4-Dinitrophenol
2-Methyl-4,6-dinitrophenol
2-Nitrophenol
4-Nitrophenol
Pentachlorophenol
Phenol
2 , 4 , 6-Tri chl orophenol
Accuracy, as
recovery, x'
(M9/L)
0.78C-3.10
1.12C+1.41
0.76C+1.58
1.09C-3.05
1.12C-6.22
0.81C-10.86
0.87C+0.06
0.84C-0.16
0.94C-0.79
0.84C+0.35
0.78C+0.29
0.87C-0.13
0.71C+4.41
0.81C-18.04
1.04C-28.04
0.07C-1.15
0.61C-1.22
0.93C+1.99
0.43C+1.26
0.91C-0.18
Single analyst
precision, sr'
(M9/L)
0.29X+1.46
0.27X+0.77
0.21X-0.41
0.19X+0.92
0.27X+0.68
0.35X+3.61
0.12X+0.57
0.16X+0.06
0.15X+0.85
0.23X+0.75
O.lSx+1.46
0.15X+1.25
0.16X+1.21
0.38X+2.36
O.lOx+42.29
0.16X+1.94
0.38X+2.57
0.24X+3.03
0.26x+0.73
0.16X+2.22
Overall
precision,
S' (M9/L)
O.BOx-0.44
0.33X+0.26
0.30X-0.68
0.27X+0.21
0.44X+0.47
0.43X+1.82
O.lBx+0.25
0.15X+0.31
0.21X+0.39
0.29X+1.31
0.28X+0.97
0.21X+1.28
0.22X+1.31
0.42X+26.29
0.26x+23.10
0.27X+2.60
0.44x+3.24
0.30X+4.33
0.35X+0.58
0.22X+1.81
X'
S'


C

x
Expected  recovery  for  one   or  more  measurements  of  a  sample
containing a concentration of C,  in M9/L-

Expected single  analyst  standard deviation of measurements  at  an
average concentration of x,  in /xg/L.

Expected interlaboratory standard deviation of measurements  at  an
average concentration found of x, in M9/L.

True value for the concentration, in ng/l.

Average recovery  found for measurements  of samples  containing  a
concentration of C, in
                                  8250A - 33
                                                        Revision 1
                                                    September 1994

-------
                                   TABLE 8.
      SURROGATE  SPIKE  RECOVERY  LIMITS  FOR  WATER  AND  SOIL/SEDIMENT SAMPLES
                                    Low/Medium             Low/Medium
Surrogate Compound                     Water              Soil/Sediment
Nitrobenzene-d5                       35-114                 23-120
2-Fluorobiphenyl                      43-116                 30-115
Terphenyl-d14                          33-141                 18-137

Phenol-d6                             10-94                  24-113
2-Fluorophenol                        21-100                 25-121
2,4,6-Tribromophenol                  10-123                 19-122
                                  8250A - 34                        Revision 1
                                                                September 1994

-------
                                        METHOD 8250A
SEMIVOLATILE  ORGANIC COMPOUNDS  BY GAS  CHROMATOGRAPHY/MASS SPECTROMETRY  (GC/MS)
                 7.1 Prepare tampla
                u*ing Method 3540,
                  3541. or 3550
7.1 Prepare (ample
u»mg Method 3510
     or 3520
                                            7.1 Prepare eample
                                            uiing Method 3540,
                                           3541, 3550, or 3580
                                               7.2 Cleanup
                                                 extract
                                                  7.3
                                              Recommended
                                                 GC/MS
                                                operating
                                                condition*.
                                                  7.4
                                                 Initial
                                               Calibration.
                                                 7.5 Daily
                                             calibration • Tune
                                             GC/MS with TFTPP
                                             and check SPCC &
                                                  CCC.
                                        8250A  -  35
          Revision 1
     September  1994

-------
        METHOD  8250A
          continued
 7.6.1 Screen extract
in GC/FID or GC/PID to
  eliminate too high
   concentration*.
     7.6.2 Spike
     sample with
       internal
      standard.
     7.6.3 Analyze
   extract by GC/MS
  using recommended
 column and operating
      conditions.
     7.7.1 Identify
     compounds by
   comparing sample
   retention time and
  sample mass spectra
     to standard*.
        7.6.4
        Does
   response exceed
   initial calibration
        curve
        range?
7.6.4 Dilute
  extract.
         7.7.2
       Quantitata
     samples using
      internal std.
       technique.
     7.7.2.4 Report
         results.
     (    Stop   J
        8250A  - 36
                  Revision 1
            September  1994

-------
oo
ks>
o\
o

-------
                                  METHOD 8260
  VOLATILE  ORGANIC  COMPOUNDS  BY  GAS  CHROMATOGRAPHY/MASS  SPECTROMETRY  (GC/MS)
                          CAPILLARY COLUMN TECHNIQUE
1.0   SCOPE AND APPLICATION

      1.1   Method 8260  is  used  to determine volatile  organic  compounds  in a
variety of solid waste matrices.   This method  is applicable to nearly all types
of samples,  regardless of water content, including ground water, aqueous sludges,
caustic  liquors,  acid liquors,  waste solvents,  oily wastes,  mousses,  tars,
fibrous  wastes,  polymeric   emulsions,   filter  cakes,  spent  carbons,  spent
catalysts, soils, and sediments.   The following compounds can be determined by
this method:
Analyte
CAS No.
  Appropriate Technique
                  Direct
Purge-and-Trap    Injection
Benzene
Bromobenzene
Bromochloromethane
Bromodichloromethane
Bromoform
Bromomethane
n-Butyl benzene
sec-Butyl benzene
tert-Butyl benzene
Carbon tetrachloride
Chlorobenzene
Chi orodi bromomethane
Chloroethane
Chloroform
Chloromethane
2-Chlorotoluene
4-Chlorotoluene
l,2-Dibromo-3-chloropropane
1,2-Dibromoethane
Di bromomethane
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
Di chl orodi f 1 uoromethane
1,1-Dichloroethane
1,2-Dichloroethane
1,1-Dichloroethene
cis-l,2-Dichloroethene
trans-l,2-Dichloroethene
1,2-Dichloropropane
1,3-Dichloropropane
71-43-2
108-86-1
74-97-5
75-27-4
75-25-2
74-83-9
104-51-8
135-98-8
98-06-6
56-23-5
108-90-7
124-48-1
75-00-3
67-66-3
74-87-3
95-49-8
106-43-4
96-12-8
106-93-4
74-95-3
95-50-1
541-73-1
106-46-7
75-71-8
75-34-3
107-06-2
75-35-4
156-59-2
156-60-5
78-87-5
142-28-9
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
PP
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
                                   8260 - 1
                                Revision 0
                                 July 1992

-------
                                     Appropriate Technique

Analyte
2,2-Dichloropropane
1,1-Dichloropropene
Ethyl benzene
Hexachlorobutadiene
Isopropyl benzene
p-Isopropyltoluene
Methyl ene chloride
Naphthalene
n-Propylbenzene
Styrene
1,1,1 , 2-Tetrachl oroethane
1 , 1 ,2, 2-Tetrachl oroethane
Tetrachloroethene
Toluene
1,2, 3 -Tri chl orobenzene
1 , 2 , 4 -Tri chl orobenzene
1,1,1-Trichloroethane
1,1,2-Trichloroethane
Trichloroethene
Tri chl orofl uoromethane
1,2,3-Trichloropropane
1 , 2 , 4-Tri methyl benzene
1 ,3,5-Trimethyl benzene
Vinyl chloride
o-Xylene
m-Xylene
p-Xylene
a Adequate response by thi

CAS No.b
594-20-7
563-58-6
100-41-4
87-68-3
98-82-8
99-87-6
75-09-2
91-20-3
103-65-1
100-42-5
630-20-6
79-34-5
127-18-4
108-88-3
87-61-6
120-82-1
71-55-6
79-00-5
79-01-6
75-69-4
96-18-4
95-63-6
108-67-8
75-01-4
95-47-6
108-38-3
106-42-3
s technique.

Purge-and-Trap
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a

Direct
Injection
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a

b Chemical Abstract Services Registry Number.
pp Poor purging efficiency
i Inappropriate technique
resulting in high
for this analyte.
EQLs.



pc Poor chromatographic behavior.
     1.2   Method 8260 can be used to quantitate most volatile organic compounds
that have boiling points below 200°C  and that are insoluble or slightly soluble
in water.  Volatile water-soluble compounds can  be included in this analytical
technique.  However,  for the more soluble compounds,  quantitation  limits are
approximately  ten times  higher  because  of poor  purging  efficiency.    Such
compounds  include low-molecular-weight  halogenated  hydrocarbons,  aromatics,
ketones, nitriles, acetates,  acrylates,  ethers, and sulfides.  See Tables 1 and
2 for lists of  analytes  and retention times that have been evaluated on a purge-
and-trap  GC/MS  system.   Also,  the  method detection  limits  for 25  ml  sample
volumes are presented.
                                   8260 - 2
Revision 0
 July 1992

-------
      1.3   The  estimated  quantitation  limit   (EQL)  of  Method  8260  for  an
 individual  compound  is  approximately  5  M9A9  (wgt  weight)  for  soil/sediment
 samples,  0.5  mg/kg  (wet weight) for wastes, and  5  ^g/l for ground water (see
 Table 3).  EQLs will be  proportionately  higher  for  sample extracts  and  samples
 that require dilution or reduced sample size to avoid saturation of the  detector.

      1.4   Method 8260 is based upon a  purge-and-trap, gas chromatographic/mass
 spectrometric  (GC/MS) procedure.  This method is restricted  to  use by, or under
 the supervision of, analysts experienced in the use of purge-and-trap systems and
 gas chromatograph/mass spectrometers, and skilled in the interpretation  of mass
 spectra and their use as a quantitative  tool.


 2.0   SUMMARY OF METHOD

      2.1   The volatile  compounds are  introduced into the gas  chromatograph by
•the  purge-and-trap  method or  by  direct  injection  (in  limited applications).
 Purged  sample components  are  trapped in  a tube containing  suitable  sorbent
 materials.  When purging is complete, the sorbent tube is heated and  backflushed
 with  helium  to desorb  trapped  sample  components.  The analytes  are desorbed
 directly  to  a large bore  capillary or  cryofocussed on a  capillary  precolumn
 before  being  flash  evaporated  to a narrow bore capillary  for analysis.   The
 column is temperature programmed to separate the  analytes which are then detected
 with  a  mass  spectrometer (MS)  interfaced to the gas chromatograph.   Wide bore
 capillary columns require a jet  separator, whereas narrow bore capillary  columns
 can be directly interfaced to the ion  source.

      2.2   If  the  above sample introduction techniques are  not  applicable,  a
 portion of the sample is dispersed  in  solvent to dissolve the  volatile  organic
 constituents.  A portion of the solution  is combined with organic-free  reagent
 water  in the  purge  chamber.   It  is  then  analyzed by  purge-and-trap GC/MS
 following the  normal water method.

      2.3   Qualitative identifications  are confirmed by analyzing standards under
 the same  conditions used for  samples and  comparing resultant mass spectra and GC
 retention times.   Each  identified component is  quantitated by relating the MS
 response  for  an  appropriate selected  ion  produced  by  that  compound to the MS
 response  for  another ion produced by an  internal  standard.


 3.0   INTERFERENCES

      3.1   Major contaminant sources are volatile materials in the laboratory and
 impurities  in  the  inert purging gas and  in the  sorbent trap.   The  use  of non-
 polytetrafluoroethylene (PTFE)   thread   sealants,  plastic  tubing,  or  flow
 controllers with rubber components should be avoided since such materials  out-gas
.organic  compounds which  will   be concentrated  in  the  trap during the purge
 operation.  Analyses of reagent blanks  provide information about the presence of
 contaminants.  When potential  interfering peaks  are noted in blanks,  the  analyst
 should  change  the purge gas source and regenerate the molecular sieve  purge gas
 filter  (Figure  1).    Subtracting  blank  values  from  sample  results  is  not
 permitted.   If reporting  values  not corrected  for blanks  result in what the
 laboratory feels is a false positive for a sample, this should be fully  explained
 in  text  accompanying the uncorrected data.

                                    8260  -  3                         Revision  0
                                                                      July 1992

-------
     3.2   Interfering contamination  may  occur when  a sample  containing  low
concentrations of  volatile  organic  compounds is analyzed  immediately  after a
sample  containing  high  concentrations  of  volatile organic  compounds.   The
preventive technique  is rinsing of the  purging apparatus and sample syringes with
two portions of organic-free reagent water between samples.  After analysis of
a sample containing  high concentrations of  volatile organic compounds,  one or
more calibration  blanks should be analyzed  to check for cross contamination.  For
samples containing large amounts of water soluble materials, suspended solids,
high boiling compounds or high concentrations of compounds being determined, it
may be necessary  to wash  the purging device with a soap solution, rinse it with
organic-free reagent  water,  and  then dry the purging device in  an oven at 105°C.
In extreme situations, the whole purge and trap device may require dismantling
and cleaning.  Screening of the samples  prior to purge and trap GC/MS analysis
is highly recommended to prevent contamination of the system. This is especially
true for soil and waste samples.  Screening may be accomplished with an automated
headspace technique  or by Method  3820 (Hexadecane  Extraction  and  Screening of
Purgeable Organics).

     3.3   Special  precautions  must  be taken to analyze for methylene chloride.
The analytical and sample storage area should be isolated from all  atmospheric
sources of methylene  chloride.   Otherwise  random background levels will  result.
Since  methylene   chloride   will   permeate   through  PTFE  tubing,   all   gas
chromatography carrier gas  lines  and  purge  gas  plumbing  should be constructed
from stainless steel  or copper  tubing. Laboratory clothing worn by the analyst
should be clean  since clothing  previously exposed  to methylene chloride fumes
during   liquid/liquid  extraction   procedures   can   contribute  to   sample
contamination.

     3.4   Samples can  be   contaminated  by  diffusion of volatile  organics
(particularly methylene chloride and fluorocarbons) through the septum seal into
the sample during shipment and storage.  A  trip blank prepared from organic-free
reagent water and carried through the sampling and handling protocol  can serve
as a check on such contamination.
4.0  APPARATUS AND MATERIALS

     4.1   Purge-and-trap device - The purge-and-trap device consists of three
separate pieces of  equipment:  the sample purger, the trap,  and  the desorber.
Several complete devices are commercially available.

           4.1.1  The recommended purging chamber is designed to accept 5 ml (and
     25 mL  if the  lowest  detection  limit  is  required)  samples with  a  water
     column at least 3 cm deep.  The gaseous-headspace between the water column
     and the trap must have a total volume  of  less  than  15  mL.   The purge gas
     must  pass  through  the water  column  as  finely divided  bubbles with  a
     diameter of less than 3 mm  at the origin.   The purge  gas must be introduced
     no more than 5  mm from the base  of  the water column.   The sample purger,
     illustrated in  Figure 1,  meets  these  design criteria.   Alternate sample
     purge devices (i.e.  needle  spargers), may  be utilized, provided equivalent
     performance is demonstrated.

           4.1.2  The trap  must be at  least   25  cm long  and have  an  inside
     diameter of  at least 0.105 in.   Starting from the inlet,  the trap must

                                   8260 - 4                         Revision 0
                                                                     July 1992

-------
     contain the following amounts of adsorbents:  1/3 of 2,6-diphenylene oxide
     polymer,  1/3  of silica gel, and 1/3 of coconut  charcoal.   It is recommended
     that 1.0  cm of methyl silicone-coated packing  be inserted at the inlet to
     extend the life of  the  trap (see Figure  2).   If it  is  not  necessary to
     analyze  for  dichlorodifluoromethane  or other  fluorocarbons of  similar
     volatility, the charcoal  can  be eliminated and the  polymer  increased to
     fill 2/3  of  the  trap.   If only  compounds  boiling  above 35°C  are  to be
     analyzed,  both the   silica  gel  and  charcoal  can be  eliminated and  the
     polymer increased  to fill the entire  trap.  Before  initial  use, the trap
     should be conditioned overnight  at 180°C by backflushing with an inert gas
     flow of at least 20 mL/min.  Vent the trap  effluent to  the room, not to the
     analytical column.   Prior to daily use,  the trap should be conditioned for
     10 minutes at  180°C  with backflushing.   The  trap  may  be vented  to  the
     analytical column  during daily  conditioning;  however, the  column must be
     run through the temperature program  prior  to  analysis of samples.   Traps
     normally  last 2-3  months when used daily.  Some  signs of a deteriorating
     trap are:  uncharacteristic recoveries of surrogates,  especially toluene-d-;
     a loss of the response of the internal  standards during  a  12 hour  shift;
     and/or a  rise in the baseline in the early portion  of the scan.

           4.1.3  The desorber  should be capable of rapidly heating the trap to
     180°C  for desorption.   The trap  bake-out  temperature should not  exceed
     220°C.  The desorber  design  illustrated  in  Figure 2  meets these  criteria.

           4.1.4  The purge-and-trap  device may be assembled as a separate unit
     or may be coupled  to a gas chromatograph,  as  shown  in Figures 3  and 4.

           4.1.5  Trap  Packing  Materials

                  4.1.5.1     2,6-Diphenylene   oxide   polymer   -   60/80   mesh,
           chromatographic grade (Tenax GC or equivalent).

                  4.1.5.2     Methyl  silicone packing - OV-1 (3%) on Chromosorb-
           W,  60/80 mesh  or equivalent.

                  4.1.5.3     Silica  gel   -  35/60  mesh,   Davison,  grade  15 or
           equivalent.

                  4.1.5.4     Coconut charcoal  - Prepare  from  Barnebey Cheney,
           CA-580-26 lot  #M-2649 by  crushing   through  a  26  mesh screen  (or
           equivalent).

     4.2   Heater  or heated  oil  bath  -  Should  be  capable of  maintaining  the
purging chamber to within 1°C over the  temperature  range  of ambient  to  100°C.

     4.3   Gas chromatography/mass spectrometer/data system

           4.3.1  Gas  chromatograph  -  An  analytical  system  complete  with  a
     temperature-programmable gas chromatograph suitable  for splitless injection
     and all required  accessories, including syringes, analytical  columns, and
     gases. The GC should be equipped  with variable constant differential flow
     controllers so that  the column  flow  rate  will  remain constant throughout
     desorption   and   temperature  program  operation.     For   some   column
     configuration, the  column oven must be  cooled  to  < 30°C,  therefore,  a

                                   8260 - 5                         Revision 0
                                                                     July 1992

-------
subambient oven controller may be required.  The capillary column  should be
directly coupled to the source.

             4.3.1.1     Capillary precolumn interface when using cryogenic
      cooling - This  device interfaces the purge and trap concentrator to
      the  capillary  gas  chromatograph.    The  interface condenses  the
      desorbed sample components and focuses  them into  a  narrow band on an
      uncoated fused silica  capillary precolumn.  When  the  interface is
      flash heated, the sample is transferred to the analytical  capillary
      column.

                   4.3.1.1.1   During    the    cryofocussing   step,    the
             temperature of the fused silica in the  interface is maintained
             at -150°C  under  a  stream of  liquid  nitrogen.   After  the
             desorption period,   the interface must  be capable  of  rapid
             heating  to  250°C   in   15  seconds or  less  to complete  the
             transfer of analytes.

      4.3.2  Gas  chromatographic columns (Recommended)

             4.3.2.1     Column  1 -  60 m x  0.75 mm  ID  capillary column
      coated with VOCOL (Supelco),  1.5 urn film thickness,  or equivalent.

             4.3.2.2     Column  2 -  30 m x  0.53 mm  ID  capillary column
      coated with  DB-624  (J&W Scientific)  or VOCOL  (Supelco), 3  ^m film.
      thickness,  or equivalent.

             4.3.2.3     Column  3 -  30 m x  0.32 mm  ID  capillary column
      coated with  DB-5 (J&W Scientific)  or  SE-54  (Supelco),  1   urn film
      thickness,  or equivalent.

      4.3.3  Mass  spectrometer - Capable  of  scanning from 35 to  300  amu
every  2  sec or  less,  using 70  volts  (nominal)  electron energy  in  the
electron impact ionization mode. The mass spectrometer must be capable of
producing a mass spectrum for Bromofluorobenzene (BFB) which meets all of
the criteria in Table  4 when 50  ng  of the GC/MS  tuning standard  (BFB) is
injected through the  GC.  To ensure sufficient precision of mass  spectral
data,  the  desirable  MS  scan rate  allows acquisition  of at least five
spectra while a sample component elutes from the GC.

      4.3.4  GC/MS  interface  - The  GC is interfaced to the MS with an all
glass enrichment device and an all glass transfer line, but any enrichment
device or  transfer line  can  be used if the  performance specifications
described  in  Section  8.2  can be achieved.    Any GC-to-MS interface that
gives acceptable calibration  points  at  50 ng or less per injection for each
of the analytes and achieves all  acceptable performance  criteria (see Table
4) may be  used.  GC-to-MS  interfaces  constructed entirely of glass or of
glass-lined  materials   are  recommended.   Glass  can  be  deactivated  by
silanizing with dichlorodimethylsilane.  This  interface  is only needed for
the wide bore columns (> 0.53 mm ID).

      4.3.5  Data  system - A  computer system that allows the continuous
acquisition  and  storage on  machine-readable media  of all  mass  spectra
obtained throughout the duration of the chromatographic  program  must be

                              8260   - 6                         Revision 0
                                                                July 1992

-------
     interfaced to the mass  spectrometer.   The  computer must have software that
     allows  searching  any GC/MS data  file for  ions  of a  specified  mass and
     plotting such ion abundances versus time or scan number.  This  type of plot
     is defined as an Extracted  Ion  Current  Profile  (EICP).  Software must also
     be available  that  allows  integrating the abundances  in  any  EICP between
     specified  time  or scan-number  limits.    The most  recent version  of the
     EPA/NIST Mass Spectral  Library should also be available.

     4.5   Microsyringes - 10,  25,  100,  250,  500,  and 1,000 nl.

     4.6   Syringe valve - Two-way,  with Luer ends (three each),  if applicable
to the purging device.

     4.7   Syringes - 5, 10,  or  25  mL,  gas-tight  with shutoff  valve.

     4.8   Balance -  Analytical,  0.0001  g, and  top-loading,  0.1 g.

     4.9   Glass scintillation vials - 20 mL,  with  Teflon  lined  screw-caps  or
glass culture tubes with Teflon  lined screw-caps.

     4.10  Vials - 2  mL, for GC  autosampler.

     4.11  Disposable pipets -  Pasteur.

     4.12         Volumetric  flasks, Class A - 10 mL  and  100  ml,  with ground-
glass stoppers.

     4.13         Spatula  -  Stainless steel.


5.0  REAGENTS

     5.1   Reagent grade inorganic chemicals shall  be used in all tests.  Unless
otherwise indicated,  it  is intended that all  inorganic reagents  shall conform to
the  specifications of  the  Committee on  Analytical  Reagents  of the  American
Chemical Society,  where such  specifications are available.  Other grades may be
used, provided it  is  first ascertained that the reagent is of sufficiently high
purity to permit its  use without lessening the accuracy of the determination.

     5.2   Organic-free reagent  water -  All  references to water in this method
refer to organic-free reagent water, as defined in Chapter One.

     5.3   Methanol,  CH3OH - Pesticide quality or equivalent, demonstrated to be
free of analytes.   Store apart from other solvents.

     5.4   Reagent Tetraglyme -  Reagent tetraglyme is defined  as  tetraglyme in
which interference is not observed at the method detection limit of compounds of
interest.

     CAUTION:     Glycol ethers are suspected carcinogens.  All solvent handling
                  should be  done  in a  hood  while  using  proper  protective
                  equipment  to minimize  exposure to  liquid  and  vapor.

           5.4.1  Tetraglyme  (tetraethylene glycol dimethyl ether, Aldrich #17,

                                   8260 - 7                         Revision 0
                                                                     July 1992

-------
     240-5 or equivalent), C8H,aCL -  Purify by treatment at reduced pressure in
     a rotary evaporator.  The telraglyme should have a peroxide  content of less
     than 5 ppm as  indicated  by  EM Quant Test Strips  (available  from Scientific
     Products Co.,  Catalog No.  P1126-8 or equivalent).

                 5.4,1.1     Peroxides may be removed by passing the tetraglyme
           through  a column  of  activated alumina.  The tetraglyme is placed in
           a round  bottom flask  equipped with  a  standard  taper  joint,  and the
           flask is affixed to a rotary  evaporator.   The flask is immersed in a
           water bath at 90-100°C and  a  vacuum  is maintained  at  < 10 mm Hg for
           at least two hours using  a two-stage mechanical pump.   The vacuum
           system is equipped with an all-glass trap, which is maintained in a
           dry ice/methanol  bath.  Cool the  tetraglyme to ambient temperature
           and add  100 mg/L of 2,6-di-tert-butyl-4-methyl-phenol  to  prevent
           peroxide formation.   Store  the tetraglyme  in a tightly sealed screw-
           cap bottle in an  area that  is not  contaminated  by  solvent vapors.

           5.4.2 In order to demonstrate that all  interfering  volatiles have
     been removed from the tetraglyme, an organic-free reagent water/tetraglyme
     blank must be  analyzed.

     5.5   Polyethylene glycol,   H(OCH2CH2)nOH -  Free of  interferences  at  the
detection limit of  the target analytes.

     5.6   Hydrochloric acid  (1:1  v/v), HC1  -  Carefully add a measured volume
of concentrated HC1 to an equal  volume of organic-free reagent water.

     5.7   Stock solutions -  Stock solutions  may  be prepared from pure standard
materials or purchased as  certified solutions.  Prepare stock standard solutions
in methanol, using  assayed liquids or gases,  as appropriate.

           5.7.1 Place about 9.8 ml of  methanol  in a  10 mL tared ground-glass-
     stoppered volumetric flask.  Allow the flask  to stand,  unstoppered,  for
     about 10 minutes  or  until   all alcohol-wetted  surfaces have dried.  Weigh
     the flask to the nearest 0.0001  g.

           5.7.2 Add the  assayed reference material, as described below.

                 5.7.2.1     Liquids  -  Using a 100 /iL  syringe,  immediately add
           two or more drops of assayed reference  material  to the flask; then
           reweigh.   The  liquid must fall  directly  into  the  alcohol  without
           contacting the neck  of the flask.

                 5.7.2.2     Gases - To  prepare standards for any compounds that
           boil below 30°C (e.g. bromomethane,  chloroethane,  chloromethane, or
           vinyl chloride),  fill a  5  ml  valved  gas-tight  syringe with  the
           reference  standard to the  5.0 ml mark.   Lower the  needle  to 5 mm
           above the methanol meniscus.  Slowly  introduce the reference standard
           above the surface  of  the liquid.  The heavy gas  will rapidly dissolve
           in the methanol.   Standards may also be prepared by  using a lecture
           bottle  equipped  with a  Hamilton Lecture Bottle  Septum (#86600).
           Attach Teflon tubing  to the side arm relief valve and  direct  a gentle
           stream of gas  into the methanol  meniscus.


                                   8260 - 8                         Revision 0
                                                                     July 1992

-------
           5.7.3  Reweigh,  dilute  to  volume,  stopper, and then mix by inverting
     the flask  several  times.   Calculate the  concentration  in  milligrams per
     liter (mg/L)  from the  net  gain  in weight.  When compound purity is assayed
     to  be  96%  or greater,  the  weight  may  be used  without   correction  to
     calculate the concentration of the stock standard.  Commercially prepared
     stock standards may be used at any concentration if they are certified by
     the manufacturer or by an independent source.

           5.7.4  Transfer  the  stock  standard  solution  into  a   bottle with  a
     Teflon lined screw-cap.  Store, with minimal headspace, at  -10°C to  -20°C
     and protect from light.

           5.7.5  Prepare  fresh standards  for gases  every two  months or sooner
     if comparison with  check standards indicates  a problem.  Reactive compounds
     such as 2-chloroethyl  vinyl ether and styrene may need to be prepared more
     frequently.   All  other standards must  be replaced after  six  months,  or
     sooner if comparison with  check standards indicates  a  problem.   Both gas
     and liquid standards must be monitored closely by comparison  to the initial
     calibration  curve  and  by comparison to QC   check  standards.   It may  be
     necessary to replace  the  standards more  frequently  if either check exceeds
     a 25% difference.

     5.8   Secondary dilution standards - Using stock standard solutions, prepare
in methanol, secondary dilution standards containing the compounds of interest,
either singly or  mixed  together.  Secondary  dilution  standards  must be stored
with minimal headspace and should be checked  frequently  for signs of degradation
or evaporation, especially  just prior to  preparing  calibration  standards from
them.  Store in a vial with no headspace for one  week only.

     5.9   Surrogate  standards  -  The  surrogates recommended  are toluene-d8,
4-bromofluorobenzene,  and  dibromofluoromethane.   Other compounds  may be used as
surrogates, depending  upon the analysis requirements.  A stock surrogate solution
in methanol  should be prepared as described  in  Section 5.7, and  a surrogate
standard spiking solution  should be prepared from the stock at a concentration
of 50-250 M9/10 ml  in methanol.  Each sample undergoing GC/MS analysis must be
spiked with 10 /nL  of  the surrogate spiking solution prior to analysis.

     5.10  Internal   standards  -  The  recommended   internal   standards  are
chlorobenzene-d5,    1,4-difluorobenzene,     1,4-dichlorobenzene-d4,    and
pentafluorobenzene.  Other compounds may be used  as  internal  standards as long
as they have retention times similar to the compounds being detected by GC/MS.
Prepare  internal  standard  stock and secondary dilution  standards  in methanol
using the procedures described in  Sections 5.7  and 5.8.   It is recommended that
the secondary dilution standard  should be prepared at a concentration of 25 mg/L
of each  internal  standard compound.   Addition of 10  /xL of this  standard  to
5.0 mL of sample or calibration standard would  be the equivalent of 50 p.g/1.

     5.11  4-Bromofluorobenzene (BFB)  standard  -  A standard solution containing
25 ng/jitL of BFB in  methanol should be prepared.

    -5.12  Calibration standards - Calibration standards  at  a  minimum of five
concentrations should be prepared from the secondary dilution of stock standards
(see Sections  5.7 and 5.8).  Prepare these  solutions  in organic-free reagent
water.  One of the concentrations  should be  at  a  concentration near, but above,

                                   8260 - 9                         Revision 0
                                                                     July 1992

-------
the method detection limit.  The remaining concentrations should correspond to
the expected range of concentrations  found in real samples but should  not exceed
the working range  of  the GC/MS system. Each standard should contain  each analyte
for detection by this method (e.g.  some  or  all  of the  compounds  listed in Table
1 may be included).  Calibration standards must  be prepared  daily.

     5.13  Matrix  spiking standards - Matrix spiking standards should be prepared
from volatile organic compounds  which will  be  representative of the compounds
being  investigated.    At  a minimum,  the  matrix spike  should  include  1,1-
dichloroethene,  trichloroethene, chlorobenzene,  toluene, and benzene.   It is
desirable  to  perform a  matrix spike using  compounds found  in  samples.   Some
permits may require  spiking specific compounds of interest, especially if they
are  polar  and would not be represented by  the  above listed compounds.   The
standard  should  be  prepared in  methanol,   with each compound present at  a
concentration of  250 /ig/10.0 ml.

     5.14  Great care must  be taken  to  maintain  the  integrity of all standard
solutions.   It  is recommended all  standards  in methanol  be  stored at -10°C to
-20°C in amber bottles with Teflon  lined screw-caps.


6.0  SAMPLE" COLLECTION,  PRESERVATION, AND HANDLING

     6.1   See  the  introductory  material to this chapter,  Organic  Analytes,
Section 4.1.
7.0  PROCEDURE

     7.1   Direct injection -  In very limited applications (e.g. aqueous process
wastes) direct  injection of  the  sample  into  the GC/MS  system  with a  10  /iL
syringe may  be  appropriate.   One such  application  is  for verification  of the
alcohol content  of an  aqueous  sample prior to  determining if  the  sample  is
ignitable (Methods  1010  or 1020).   In this case,  it  is  suggested that  direct
injection  be  used.     The   detection   limit   is  very  high   (approximately
10,000 M9/L).  Therefore, it is only permitted  when  concentrations  in excess of
10,000 p.g/1 are expected, or for water-soluble compounds that do not purge.  The
system must be calibrated by direct injection using the same solvent (e.g. water)
for standards as the sample matrix  (bypassing the purge-and-trap device).

     7.2   Chromatographic conditions (Recommended)

           7.2.1  General:

           Injector temperature:           200-225°C
           Transfer line temperature:      250-300°C

           7.2.2  Column 1  (A sample chromatogram is presented in Figure 5)

           Carrier gas (He) flow rate:    15 mL/min
           Initial temperature:           10°C, hold for 5 minutes
           Temperature program:           6°C/min to 160°C
           Final temperature:             160°C,   hold  until   all   expected
                                          compounds have  eluted.

                                   8260  - 10                        Revision 0
                                                                     July 1992

-------
      7.2.3   Column 2, Cryogenic cooling  (A sample chromatogram is presented
in Figure 6)
      Carrier gas  (He)  flow rate:
      Initial  temperature:
      Temperature  program:
      Final  temperature:
              15  mL/min
              10°C, hold  for 5 minutes
              6°C/min  to  160°C
              160°C,   hold  until   all
              compounds  have eluted.
                            expected
      7.2.4  Column  2,
presented in Figure 7)
Non-cryogenic  cooling  (A  sample  chromatogram  is
      Carrier gas flow rate:
      Initial  temperature:
      Temperature program:
      Final  temperature:
       It  is  recommended  that  carrier  gas  flow  and
       split   and  make-up  gases  be  set  using
       performance of standards as guidance.   Set
       the carrier gas head pressure  to'» 10  psi
       and the split  to  = 30 mL/min.   Optimize  the
       make-up   gas   flow   for   the  separator
       (approximately 30  mL/min)  by injecting BFB,
       and determining the  optimum response when
       varying the make-up gas.   This  will require
       several   injections of  BFB.    Next,  make
       several  injections of the volatile working
       standard  with  all  analytes  of interest.
       Adjust  the  carrier  and split  to  provide
       optimum chromatography  and  response.  This
       is  an  especially critical adjustment for  the
       volatile gas analytes.   The head  pressure
       should  optimize  between 8-12  psi  and  the
       split  between  20-60 mL/min.   The use  of  the
       splitter is  important to minimize the  effect
       of  water on analyte response,  to allow  the
       use of a larger volume of helium during trap
       desorption,  and to slow column flow.
       45°C,  hold for  2 minutes
       8°C/min  to 200°C
       200°C,  hold  for 6  minutes.
      A trap preheated  to  150°C  prior to trap desorption  is  required to
provide adequate chromatography of the gas analytes.

      7.2.5  Column  3  (A sample chromatogram is  presented in Figure 8)
      Carrier gas (He)  flow rate:
      Initial temperature:
      Temperature program:

      Final  temperature:
                                                to
4 mL/min
10°C,  hold  for 5  minutes
6°C/min   to  70°C,  then  15°C/min
145°C
145°C,    hold   until    all   expected
compounds have eluted.
7.3   Initial calibration for purge-and-trap procedure

      7.3.1  Each GC/MS system must  be  hardware-tuned  to meet the criteria
in Table 4 for a 50 ng injection or purging of 4-bromofluorobenzene (2 /iL
                              8260 -  11
                                        Revision 0
                                         July 1992

-------
 injection of  the BF3  standard).    Analyses Tiust  not begin  until  these
 criteria are met.

       7.3.2   Assemble  a  purge-and-trap device that meets the specification
 in Section 4.1.  Condition  the  trap overnight  at 1SO°C in the purge mode
 with an inert gas flow of  at least  20 mL/min.   Prior to use, condition the
 trap daily for 10 minutes while  backflushing  at 180°C with the column at
 220bC.

       7.3.3   Connect the purge-and-trap  device  to a  gas chromatograph.

       7.3.4   A set  of  at least five  calibration standards containing the
 method analytes  is needed.   One  calibration standard  should contain each
 analyte  at   a concentration  approaching  but   greater than  the  method
 detection  limit   (Table  1)  for  that   compound; the other  calibration
 standards should  contain analytes at concentrations that define the range
 of the method.   The purging  efficiency  for  5 mL of  water  is greater than
 for 25 ml.   Therefore, develop the  standard  curve with whichever volume of
 sample that will  be analyzed.  To  prepare  a calibration  standard,  add an
 appropriate  volume  of  a  secondary dilution standard solution to an aliquot
 of organic-free  reagent  water in a  volumetric  flask.   Use a microsyringe
 and rapidly inject the  alcoholic  standard  into the expanded  area  of the
 filled volumetric flask.  Remove the needle as quickly as possible after
 injection.    Mix  by inverting the  flask three  times  only.   Discard the
 contents contained in the neck  of the  flask.   Aqueous standards are not
 stable and should be prepared daily.  Transfer 5.0  ml (or 25 ml if lower
 detection limits are  required)  of  each  standard to  a gas  tight syringe
 along with 10 /ul_ of  internal standard.   Then  transfer the contents to a
 purging device.

       7.3.5   Carry  out the purge-and-trap analysis procedure as described
 in Section 7.5.1.

       7.3.6   Tabulate  the  area response  of the  characteristic  ions (see
 Table  5) against  concentration  for  each  compound   and  each  internal
 standard.  Calculate response factors (RF)  for each  compound relative to
 one of  the  internal  standards.    The  internal   standard selected for the
 calculation of the RF  for  a compound should be  the internal standard that
 has  a  retention  time  closest to  the   compound being measured  (Section
 7.6.2).  The RF  is calculated as  follows:

                          RF  = (AXC,S)/(AJSCX)

where:

       Ax     =     Area  of the characteristic  ion for the compound being
                    measured.
       Ais     =     Area  of the characteristic ion for the specific internal
                    standard.
       Cis     =     Concentration  of the  specific internal  standard.
       Cx     =     Concentration  of the  compound being measured.

       7.3.7  The  average  RF must  be calculated and recorded  for each
 compound.    A  system  performance  check   should   be made  before  this

                               8260  - 12                         Revision 0
                                                                 July 1992

-------
calibration curve  is used.   Five  compounds (the System Performance Check
Compounds, or  SPCCs)  are  checked for a  minimum average response factor.
These compounds are chloromethane; 1,1-dichloroethane; bromoform; 1,1,2,2-
tetrachloroethane; and chlorobenzene.   The minimum acceptable average RF
for these compounds should be 0.300 (0.250 for bromoform).  These compounds
typically have RFs of 0.4-0.6  and are  used to check compound instability
and to check for degradation caused by contaminated lines or active sites
in the system.  Examples of these occurrences are:

             7.3.7.1     Chloromethane  -  This  compound  is  the most likely
      compound to be lost  if the purge  flow is too fast.

             7.3.7.2    Bromoform -  This compound  is one of the compounds
      most likely to be purged very  poorly if the purge flow is too slow.
      Cold spots and/or active  sites in  the  transfer  lines  may adversely
      affect  response.   Response of the  quantitation  ion  (m/z  173)  is
      directly  affected  by  the  tuning  of  8FB  at  ions  m/z  174/176.
      Increasing  the  m/z  174/176 ratio  relative  to  m/z  95  may improve
      bromoform response.

             7.3.7.3    Tetrachloroethane  and 1,1-dichloroethane -  These
      compounds are degraded by contaminated transfer lines in purge-and-
      trap systems and/or  active sites  in trapping materials.

      7.3.8  Using the  RFs  from the  initial  calibration,  calculate  the
percent relative standard  deviation  (%RSD)  for Calibration Check Compounds
(CCCs).  Record the %RSDs for all compounds.  The percent RSD is calculated
as follows:

                                 SD
                          %RSD  = 	 x  100
where:
      RSD    =     Relative standard deviation.
      x      =     Mean  of 5 initial  RFs  for a  compound.
      SD     =     Standard deviation of  average  RFs  for  a  compound.
                         SD
N  (x,.  -  x)2
£  	
1-1  N  -  1
      The %RSD for each individual  CCC  must  be  less than 30 percent.  This
criterion must be met for the individual calibration to be  valid.  The CCCs
are:

      1,1-Dichloroethene,
      Chloroform,
      1,2-Dichloropropane,
      Toluene,
      Ethyl benzene, and
      Vinyl chloride.

                              8260  - 13                        Revision 0
                                                                July 1992

-------
7.4   DaHy GC/MS cal i brat ion

      7.4.1  Prior to the  analysis of samples, inject or purge 50  ng of the
4-bromofluorobenzene standard.   The resultant mass spectra for the BFB must
meet all of  the  criteria  given  in  Table 4 before sample analysis begins.
These criteria must be  demonstrated each 12-hour shift.

      7.4.2  The  initial calibration  curve  (Section  7.3) for each compound
of interest must  be checked  and  verified  once  every 12 hours of analysis
time.  This is accomplished  by analyzing a  calibration  standard that is at
a concentration  near the  midpoint  concentration  for the working  range of
the GC/MS by checking the SPCC (Section 7.4.3)  and CCC (Section 7.4.4).

      7.4.3  System  Performance  Check  Compounds  (SPCCs)   -  A  system
performance check must  be made  each  12 hours.    If  the SPCC  criteria are
met, a comparison of response factors is made for all compounds.   This is
the  same check that is applied  during  the initial   calibration.   If the
minimum response  factors  are not met,   the system must be  evaluated,  and
corrective action must  be  taken before sample  analysis begins.  The minimum
response factor  for volatile SPCCs is 0.300  (0.250  for Bromoform).   Some
possible problems are  standard  mixture  degradation, injection port inlet
contamination, contamination at  the front end of the  analytical column, and
active sites in the column or chromatographic system.

      7.4.4  Calibration  Check   Compounds  (CCCs)  -  After  the  system
performance check is met, CCCs  listed  in  Section 7.3.8 are used to check
the validity of the initial  calibration. Calculate  the percent difference
using the following equation:

                                 RF.  - RFC
                 % Difference = 	 x 100
                                    RF,
where:
      RF,     =     Average   response  factor   from   initial   calibration
                   (Section  7.3).
      RFC     =     Response   factor   from  current   verification   check
                   standard.

      If the  percent difference for  any  compound  is  greater than 20, the
laboratory should consider this a warning limit.  If the percent difference
for each CCC  is  less than  25%, the  initial calibration  is assumed to be
valid.  If  the criterion is  not met  (>  25% difference),  for any one CCC,
corrective  action  must  be taken.   Problems similar  to those listed under
SPCCs could  affect this criterion.   If  no source of the  problem  can be
determined  after  corrective ' action  has  been  taken,  a  new  five-point
calibration  must  be generated.    This  criterion   must  be  met  before
quantitative  sample analysis begins.   If the CCCs are  not  required analytes
by  the  permit,  then all required  analytes must meet the  25% difference
criterion.

      7.4.5  The  internal standard responses  and retention  times  in the
check calibration  standard  must be evaluated  immediately after or during

                              8260  -  14                         Revision 0
                                                                July 1992

-------
data acquisition.   If the  retention  time  for any  internal standard changes
by more than 30  seconds  from  the  last daily calibration  (Section 7.4), the
chromatographic  system must be inspected for malfunctions and corrections
must be  made,  as  required.   If  the  EICP area  for any of  the internal
standards changes  by  a  factor of two  (-50% to +100%) from the last daily
calibration standard  check,  the  mass spectrometer  must be  inspected for
malfunctions  and  corrections  must  be  made,  as  appropriate.    When
corrections are  made, reanalysis of samples analyzed while the system was
malfunctioning are necessary.

7.5   GC/MS analysis

      7.5.1  Water samples

             7.5.1.1    Screening  of the sample prior  to  purge-and-trap
      analysis  will  provide guidance   on  whether  sample  dilution  is
      necessary   and  will  prevent  contamination  of the  purge-and-trap
      system.  Two screening  techniques that can  be used are the headspace
      sampler (Method 3810) using a  gas chromatograph (GC) equipped with a
      photo  ionization  detector  (PID)   in  series  with an  electrolytic
      conductivity  detector  (HECD), and  extraction of the  sample  with
      hexadecane and analysis of the extract on a GC with a FID and/or an
      ECD (Method 3820).

             7.5.1.2    All  samples and standard solutions must be allowed
      to warm to ambient temperature before analysis.

             7.5.1.3    Set  up the GC/MS  system as outlined  in  Sections 4.3
      and 7.2.

             7.5.1.4    BFB  tuning criteria and daily  GC/MS  calibration
      criteria must be met (Section 7.4)  before analyzing samples.

             7.5.1.5    Adjust  the purge gas  (helium)  flow  rate  to 25-
      40 mL/min  on  the  purge-and-trap device.   Optimize the  flow rate to
      provide the best response for chloromethane and bromoform, if these
      compounds  are  analytes.  Excessive flow rate reduces chloromethane
      response,  whereas insufficient flow reduces bromoform response (see
      Section 7.3.7).

             7.5.1.6    Remove the  plunger from  a  5 ml  syringe and attach
      a closed syringe valve.  If lower detection limits are required, use
      a 25 ml syringe.  Open  the  sample or standard bottle, which has been
      allowed to come to ambient temperature,  and carefully pour the sample
      into the syringe  barrel to  just  short  of overflowing.   Replace the
      syringe plunger and  compress the  sample.  Open  the syringe valve and
      vent any residual air  while  adjusting  the  sample  volume to 5.0 mL.
      This process of taking  an aliquot destroys the validity of the liquid
      sample for future  analysis; therefore, if there is  only one VOA vial,
      the analyst  should  fill  a  second  syringe  at  this time to protect
      against possible  loss  of  sample integrity.   This second  sample is
      maintained only until  such  time when  the  analyst has determined that
      the  first  sample  has  been analyzed properly.    Filling one  20 ml
      syringe would allow the  use  of only  one  syringe.     If  a  second

                              8260 - 15                        Revision 0
                                                                July 1992

-------
analysis is  needed  from a  syringe,  it must be  analyzed  within 24
hours.    Care must be  taken to  prevent  air from  leaking  into the
syringe.

       7.5.1.7    The following procedure is appropriate for diluting
purgeable samples.  All  steps must  be  performed without delays until
the diluted sample is in a gas-tight syringe.

             7.5.1.7.1   Dilutions  may be made  in volumetric flasks
       (10  to 100 mL).   Select the volumetric flask that will allow
       for  the necessary  dilution.    Intermediate  dilutions  may be
       necessary  for  extremely  large dilutions.

             7.5.1.7.2   Calculate the approximate volume of organic-
       free  reagent  water  to  be   added  to  the  volumetric  flask
       selected and add  slightly less  than this quantity of organic-
       free reagent water  to the flask.

             7.5.1.7.3   Inject  the proper aliquot of sample from the
       syringe prepared  in Section  7.5.1.6 into the  flask.  Aliquots
       of less than  1 ml are not recommended.   Dilute the sample to
       the  mark  with organic-free  reagent  water.    Cap  the flask,
       invert, and shake  three  times.   Repeat  above procedure for
       additional dilutions.

             7.5.1.7.4   Fill a  5 ml syringe with the diluted sample
       as in Section  7.5.1.6.

       7.5.1.8    Compositing  samples prior to  GC/MS analysis

             7.5.1.8.1   Add 5   ml  or  equal  larger  amounts  of each
       sample (up to  5 samples are allowed) to a  25 ml glass syringe.
       Special precautions must  be made to maintain zero headspace in
       the  syringe.

             7.5.1.8.2   The samples must be  cooled  at 4°C during this
       step to minimize volatilization losses.

             7.5.1.8.3   Mix well and  draw  out  a  5 ml  aliquot for
       analysis.

             7.5.1.8.4   Follow   sample  introduction,  purging,  and
       desorption steps described in the method.

             7.5.1.8.5   If less  than  five samples  are  used  for
       compositing,   a  proportionately smaller syringe  may  be used
       unless a 25 ml sample is to be purged.

       7.5.1.9    Add   10.0  /LtL   of  surrogate  spiking  solution
(Section 5.9)  and 10   n\-  of   internal  standard  spiking  solution
(Section 5.10) through the valve bore of the syringe;  then close the
valve.  The  surrogate and  internal  standards may  be mixed and added
as a single spiking solution.  The addition of 10 /il_ of the surrogate
spiking solution to  5 ml  of sample  is  equivalent  to a concentration

                        8260 -  16                        Revision  0
                                                          July  1992

-------
of 50 ug/l of each surrogate standard.

       7.5.1.10     Attach  the  syringe-syringe valve assembly to the
syringe valve on  the  purging  device.   Open  the syringe valves and
inject the sample into the purging chamber.

       7.5.1.11     Close both valves  and purge the  sample for 11.0 ±
0.1 minutes at ambient temperature.   Be  sure  the trap  is cooler than
25°C.

       7.5.1.12     Sample  desorption  - The mode  of  sample desorption
is determined  by  the type of  capillary  column employed  for  the
analysis.    When  using  a  wide  bore  capillary  column,  follow  the
desorption conditions  of Section  7.5.1.13.  The  conditions for using
narrow bore columns are described in Section 7.5.1.14.

       7.5.1.13     Sample  desorption  for wide bore  capillary column.
Under most conditions, this type  of  column must  be  interfaced to the
MS through an all glass jet separator.

             7.5.1.13.1  After the  11 minute purge, attach the trap
       to  the chromatograph, adjust  the  purge and trap  system to the
       desorb mode (Figure 4) and  initiate  the  temperature program
       sequence  of the gas chromatograph and  start  data acquisition.
       Introduce the  trapped materials  to  the GC column by rapidly
       heating the trap to 180°C while backflushing the trap with an
       inert  gas at 15 mL/min for 4  minutes.   If the non-cryogenic
       cooling technique  is followed, the trap must be preheated to
       150°C  just prior to trap desorption at 180°C.  While the purged
       analytes   are  being introduced  into the  gas  chromatograph,
       empty  the purging device using the sample  syringe and wash the
       chamber  with  two   5  ml  or  25 ml portions of organic-free
       reagent  water  depending  on  the   size  of the  purge  device.
       After  the purging  device  has been emptied,  leave the syringe
       valve  open to allow the  purge gas to vent through the sample
       introduction needle.

             7.5.1.13.2 Hold  the  column temperature  at  10°C  for
       5 minutes, then program at 6°C/min to 160°C and  hold until all
       analytes  elute.

             7.5.1.13.3 After desorbing the sample  for  4 minutes,
       condition the trap by returning the purge-and-trap system to
       the purge mode. Wait 15  seconds,  then close  the  syringe valve
       on  the purging device  to begin  gas flow through  the trap.
       Maintain  the trap  temperature at  180°C.   After approximately
       7 minutes,  turn off the trap heater and open  the  syringe valve
       to  stop the gas flow through the trap.  When  the trap  is cool,
       the next  sample .can be  analyzed.

       7.5.1.14     Sample  desorption for  narrow bore capillary column.
Under normal  operating conditions, most narrow bore  capillary columns
can be interfaced directly to  the MS without a jet separator.
                        8260 - 17                        Revision 0
                                                          July  1992

-------
             7.5.1.14.1  After  the  11  minute purge,  attach the trap
       to the cryogenically cooled interface at -150°C  and  adjust the
       purge-and-trap   system  to  the   desorb   mode   (Figure  4).
       Introduce  the trapped materials  to  the  interface  by rapidly
       heating  the  trap  to  180°C while backflushing the trap with an
       inert  gas at  4 mL/min for 5  minutes.   While  the  extracted
       sample   is  being  introduced  into the  interface,   empty  the
       purging  device using  the sample syringe and rinse the chamber
       with  two 5 ml or  25 ml portions of organic-free reagent water
       depending  on  the size  of the  purging device.    After  the
       purging  device has been emptied,  leave the syringe valve open
       to allow the purge gas to vent  through the sample introduction
       needle.    After  desorbing for   5  minutes,  flash  heat  the
       interface  to 250°C  and  quickly  introduce  the  sample  on  the
       chromatographic   column.     Start  the  temperature  program
       sequence,  and initiate data  acquisition.

             7.5.1.14.2  Hold  the column temperature  at   10°C  for
       5 minutes, then program at 6°C/min to  70 C and then at 15°C/min
       to  145°C.    After  desorbing  the   sample  for  5   minutes,
       recondition the trap by returning  the  purge-and-trap system to
       the purge mode.  Wait 15 seconds,  then close the syringe valve
       on the  purging device to begin  gas  flow through  the trap.
       Maintain the trap temperature at 180°C.   After approximately
       15 minutes,   turn off the trap  heater  and open  the syringe
       valve  to stop the gas flow through the trap.  When the trap is
       cool,  the  next  sample can  be  analyzed.

       7.5.1.15    If the initial  analysis of sample or a dilution of
the sample has  a concentration of analytes  that exceeds the initial
calibration  range,  the sample  must  be  reanalyzed  at  a  higher
dilution.  Secondary ion quantitation  is allowed only when there are
sample interferences with the primary ion.  When a sample is analyzed
that  has saturated  ions  from a  compound, this  analysis must  be
followed  by  a  blank organic-free reagent  water  analysis.   If the
blank  analysis  is  not  free  of  interferences,  the  system must  be
decontaminated.  Sample  analysis  may  not resume until  a blank can be
analyzed that is free of interferences.

       7.5.1.16    For matrix spike analysis, add  10 /xL of  the matrix
spike  solution (Section 5.13)  to the 5 ml  of  sample  to  be purged.
Disregarding any dilutions,  this  is equivalent to a concentration of
50 jug/L  of each matrix spike standard.

       7.5.1.17    All dilutions should keep  the response of the major
constituents (previously saturated peaks)  in  the  upper  half of the
linear range of the curve.   Proceed to  Sections 7.6.1 and 7.6.2 for
qualitative and quantitative analysis.

7.5.2  Water-miscible liquids

       7.5.2.1      Water-miscible  liquids   are  analyzed  as  water
samples  after first diluting them at  least  50  fold with organic-free
reagent  water.

                        8260 -  18                         Revision  0
                                                          July 1992

-------
             7.5.2.2     Initial  and serial dilutions  can  be prepared by
      pipetting  2  ml  of  the  sample  to  a 100  ml  volumetric  flask and
      diluting  to   volume  with  organic-free  reagent  water.    Transfer
      immediately to a 5 ml gas-tight syringe.

             7.5.2.3    Alternatively,  prepare  dilutions directly  in a 5
      ml syringe filled with  organic-free  reagent water by adding at least
      20 ML,  but not more than 100  nl  of liquid sample.   The sample is
      ready for addition of internal and surrogate  standards.

      7.5.3  Sediment/soil  and waste samples -  It is highly recommended that
all  samples  of this type  be  screened  prior  to  the  purge-and-trap GC/MS
analysis.  The headspace method (Method  3810)  or the hexadecane extraction
and  screening method  (Method  3820) may  used for  this purpose.   These
samples  may  contain percent quantities of purgeable  organics  that  will
contaminate the  purge-and-trap  system,  and require  extensive cleanup and
instrument downtime.   Use  the  screening data  to  determine  whether  to use
the  low-concentration  method  (0.005-1  mg/kg) or the high-concentration
method (> 1 mg/kg).

             7.5.3.1    Low-concentration  method -  This is  designed for
      samples containing individual  purgeable compounds of < 1 mg/kg.  It
      is limited to  sediment/soil samples  and waste  that is  of a similar
      consistency (granular and porous).  The low-concentration method is
      based on purging  a heated sediment/soil  sample  mixed with organic-
      free reagent water containing the surrogate and  internal standards.
      Analyze all blanks  and  standards  under the same conditions  as the
      samples.  See Figure 9 for an illustration  of  a low soils impinger.

                   7.5.3.1.1    Use   a   5   g  sample   if  the  expected
             concentration  is  < 0.1 mg/kg or a  1  g  sample  for expected
             concentrations between  0.1  and 1  mg/kg.

                   7.5.3.1.2    The  GC/MS system  should  be  set .up as in
             Sections 7.5.1.3-7.5.1.4.   This  should be done  prior  to the
             preparation of  the  sample  to  avoid  loss  of  volatiles  from
             standards and  samples.  A heated  purge calibration curve must
             be  prepared  and  used  for  the quantitation of  all  samples
             analyzed  with the  low-concentration  method.     Follow  the
             initial and daily calibration instructions, except  for the
             addition of a 40°C purge temperature.

                   7.5.3.1.3    Remove the plunger from a 5 ml  Luerlock type
             syringe  equipped  with  a   syringe  valve  and   fill   until
             overflowing with water.  Replace  the plunger and compress the
             water to vent  trapped  air.  Adjust the volume to 5.0 ml.   Add
             10 ML  each of surrogate spiking solution (Section 5.9) and
             internal  standard solution  (Section  5.10)  to  the  syringe
             through the valve (surrogate spiking solution  and internal
             standard solution may  be mixed together).  The addition of
             10 /.iL   of  the   surrogate   spiking  solution   to  5   g  of
             sediment/soil  is  equivalent  to  50  M9A9  of  each surrogate
             standard.
                              8260 -  19                        Revision 0
                                                                July 1992

-------
      7.5.3.1.4   The  sample  (for  volatile  organics) consists
of  the  entire  contents  of  the  sample  container.    Do not
discard any  supernatant  liquids.    Mix the  contents  of the
sample  container  with a  narrow  metal  spatula.    Weigh the
amount  determined  in  Section  7.5.3.1.1  into  a  tared  purge
device.  Note and  record  the  actual  weight  to the  nearest 0.1
9-

      7.5.3.1.5   Determine  the  percent  dry weight  of the
soil/sediment sample.   This  includes  waste samples that are
amenable to  percent dry  weight determination.  Other wastes
should be reported on  a wet-weight basis.

            7.5.3.1.5.1   Immediately after weighing the sample
      for extraction,  weigh 5-10 g of  the sample into a  tared
      crucible.   Determine the % dry weight of the sample by
      drying  overnight  at   105°C.    Allow   to   cool   in  a
      desiccator  before   re-weighing.    Concentrations  of
      individual analytes are  reported relative  to  the dry
      weight of sample.

            WARNING:      The  drying  oven  should be contained
                         in  a  hood  or vented.    Significant
                         laboratory  contamination  may result
                         from a heavily contaminated hazardous-
                         waste  sample.

            % dry  weight  = g  of dry  sample  x  100
                               g of  sample

      7.5.3.1.6   Add  the spiked organic-free reagent water to
the  purging device,   which  contains  the  weighed  amount  of
sample, and connect the device to the purge-and-trap system.

      NOTE:       Prior to the attachment  of the purge device,
                  the  procedures  in  Sections  7.5.3.1.4 and
                  7.5.3.1.6  must  be  performed rapidly and
                  without  interruption  to  avoid  loss  of
                  volatile organics.   These steps  must  be
                  performed  in  a  laboratory  free  of  solvent
                  fumes.

      7.5.3.1.7   Heat the sample  to 40°C +  1°C and purge the
sample  for  11.0 + 0.1 minutes.   Be  sure  the trap is cooler
than 25°C.

      7.5.3.1.8   Proceed with  the  analysis  as  outlined  in
Sections 7.5.1.12-7.5.1.17.   Use 5 mL of the same organic-free
reagent water as in the blank.  If saturated peaks  occurred or
would  occur  if  a  1  g  sample  were analyzed,   the   high-
concentration method must be followed.

      7.5.3.1.9   For   matrix    spike   analysis   of   low-
concentration sediment/soils,  add 10  juL  of  the matrix  spike

                 8260 - 20                         Revision 0
                                                   July 1992

-------
       solution (Section 5.7)  to  the 5 ml  of organic-free reagent
       water (Section  7.5.3.1.3).  The concentration for a  5 g sample
       would be equivalent to 50 /^g/kg of each matrix  spike  standard.

       7.5.3.2     High-concentration method  - The method is based on
extracting the sediment/soil with methanol.  A waste sample  is either
extracted  or  diluted,  depending on its  solubility  in   methanol.
Wastes  (i.e.  petroleum  and  coke  wastes)  that  are  insoluble  in
methanol are diluted with tetraglyme  or possibly polyethylene glycol
(PEG).  An  aliquot  of  the extract  is added to organic-free reagent
water containing  surrogate and  internal standards.  This is  purged at
ambient temperature.  All  samples with an expected concentration of
> 1.0 mg/kg should be analyzed by this method.

             7.5.3.2.1   The sample (for volatile organics) consists
       of  the  entire  contents  of the  sample  container.   Do  not
       discard any  supernatant liquids.    Mix the contents of  the
       sample  container  with   a   narrow  metal  spatula.     For
       sediment/soil and solid wastes that  are insoluble in methanol
       weigh 4 g  (wet weight) of sample into a tared 20 ml  vial.  Use
       a top-loading balance.   Note and record the actual  weight to
       0.1 gram and determine  the percent  dry weight of the sample
       using the procedure in Section 7.5.3.1.5.  For waste that is
       soluble in  methanol,  tetraglyme,  or  PEG, weigh   1 g  (wet
       weight) into a tared scintillation vial or culture  tube or  a
       10 ml volumetric flask.  (If a vial or tube is used, it must
       be calibrated prior to use.  Pipet 10.0 ml of solvent into the
       vial  and  mark  the  bottom of the  meniscus.    Discard  this
       solvent.)

             7.5.3.2.2   For  sediment/soil  or solid  waste, quickly
       add  9.0 ml  of appropriate solvent;  then add   1.0  ml  of the
       surrogate  spiking  solution  to  the  vial.    For  a solvent
       miscible  sample,  dilute  the  sample  to   10  ml   with  the
       appropriate  solvent  after adding  1.0  ml  of  the   surrogate
       spiking solution.  Cap and shake for 2 minutes.

             NOTE:       Sections  7.5.3.2.1  and 7.5.3.2.2 must  be
                        performed  rapidly and without  interruption
                        to  avoid  loss of volatile organics.   These
                        steps must be performed in a  laboratory free
                        from  solvent fumes.

             7.5.3.2.3  Pipet  approximately  1 ml of  the extract to
       a  GC  vial   for  storage,  using  a  disposable  pipet.    The
       remainder may  be disposed.    Transfer  approximately 1  ml of
       appropriate  solvent  to a  separate GC  vial  for use  as  the
       method blank for each set of  samples.  These  extracts may be
       stored at  4°C in the dark, prior to analysis.   The addition of
       a  100  ML  aliquot  of  each   of  these  extracts  in Section
       7.5.3.2.6 will  give  a concentration  equivalent to 6,200 M9/kg
       of each surrogate standard.
                        8260 - 21                         Revision  0
                                                           July  1992

-------
                  7.5.3.2.4   The  GC/MS  system  should  be set  up as  in
            Sections  7.5.1.3-7.5.1.4.   This should be done  prior  to  the
            addition of the solvent extract  to organic-free reagent water.

                  7.5.3.2.5   The  information in Table 10 can  be  used to
            determine the volume of solvent extract to add to the 5 ml of
            organic-free  reagent  water  for  analysis.    If  a  screening
            procedure  was  followed  (Method  3810  or  3820),  use  the
            estimated concentration to determine the  appropriate  volume.
            Otherwise, estimate the concentration range of the sample from
            the  low-concentration  analysis  to determine  the appropriate
            volume.   If the sample was submitted as  a high-concentration
            sample,  start with  100 nl.    All  dilutions must keep  the
            response  of  the  major constituents  (previously  saturated
            peaks)  in the upper  half of the linear range  of  the curve.

                  7.5.3.2.6   Remove the plunger from a  5.0 ml  Luerlock
            type  syringe  equipped  with  a   syringe valve  and  fill  until
            overflowing with water.  Replace the plunger and  compress  the
            water to vent trapped air.  Adjust the volume to 4.9 ml.  Pull
            the plunger back to  5.0 ml to allow volume for the addition of
            the  sample extract  and of  standards.  Add 10 /A of  internal
            standard  solution.   Also add the volume of  solvent  extract
            determined in Section  7.5.3.2.5 and a volume of extraction or
            dissolution  solvent to total  100 /*L  (excluding  solvent  in-
            standards).

                  7.5.3.2.7   Attach the syringe-syringe valve assembly to
            the  syringe  valve on  the purging  device.  Open the  syringe
            valve  and inject  the  water/solvent  sample into the  purging
            chamber.

                  7.5.3.2.8   Proceed with  the  analysis  as  outlined  in
            Sections  7.5.1.12-7.5.1.17.   Analyze all blanks on  the same
            instrument as  that  used for the samples.  The  standards  and
            blanks  should  also  contain 100  /iL of the dilution  solvent to
            simulate  the  sample  conditions.

                  7.5.3.2.9   For a matrix spike  in the high-concentration
            sediment/soil  samples, add  8.0 ml  of methanol,  1.0 ml  of
            surrogate spike solution  (Section  5.9),  and  1.0  ml of.matrix
            spike  solution  (Section 5.13) as  in  Section  7.5.3.2.2.  This
            results in  a  6,200  /ug/kg concentration  of each  matrix spike
            standard  when added  to a-4 g sample.  Add a 100 pi aliquot of
            this extract to 5  ml of organic-free  reagent water  for purging
            (as  per Section 7.5.3.2.6).

7.6   Data interpretation

      7.6.1 Qualitative analysis

            7.6.1.1    An  analyte  (e.g.  those listed   in  Table  1)  is
      identified  by comparison of  the sample  mass  spectrum  with  the mass
      spectrum of a  standard of  the suspected compound (standard reference

                              8260  - 22                         Revision  0
                                                                July 1992

-------
spectrum).  Mass spectra  for  standard reference should be obtained on
the user's  GC/MS  within  the same  12  hours  as the sample  analysis.
These standard reference  spectra may be obtained  through analysis of
the calibration standards.   Two criteria must  be  satisfied  to verify
identification:  (1)  elution  of  sample component  at  the  same GC
relative retention time (RRT) as those of the standard component; and
(2) correspondence of the sample component and the standard  component
mass spectrum.

             7.6.1.1.1    The  sample component RRT  must compare within
       + 0.06 RRT units of the  RRT  of the standard component.   For
       reference,  the standard must be run within the same 12 hours
       as  the  sample.    If  coelution  of  interfering  components
       prohibits accurate assignment of the sample component RRT from
       the  total  ion chromatogram,  the  RRT should  be  assigned by
       using extracted ion current  profiles  for  ions unique to the
       component of interest.

             7.6.1.1.2    (1)  All  ions present  in  the standard  mass
       spectra  at a  relative  intensity  greater  than 10%  (most
       abundant ion  in the spectrum  equals  100%  must be present in
       the  sample spectrum).   (2)  The relative intensities of ions
       specified in (1) must  agree within ±  20% between  the standard
       and sample spectra.   Example:  For an  ion with  an  abundance of
       50%  in  the  standard  spectra,  the   corresponding  sample
       abundance must be  between 30 and 70 percent.

       7.6.1.2     For  samples  containing components not associated
with the calibration  standards,  a  library search  may  be  made for the
purpose of  tentative identification.  The necessity  to  perform this
type of  identification will  be determined  by  the type  of analyses
being conducted.  Guidelines for making tentative  identification are:

       (1)   Relative intensities  of major  ions  in  the  reference
spectrum  (ions >  10% of the most abundant ion) should be present in
the sample  spectrum.

       (2)   The relative intensities  of the major ions  should agree
within + 20%.  (Example:   For an ion with an abundance of 50% in the
standard  spectrum, the  corresponding sample  ion  abundance must be
between 30  and 70%).

       (3)   Molecular  ions present  in  the reference  spectrum should
be present  in the sample  spectrum.

       (4)   Ions present  in the  sample spectrum  but  not  in the
reference   spectrum  should  be reviewed for possible background
contamination or presence of coeluting compounds.

       (5)   Ions present in the  reference  spectrum  but not in the
sample spectrum should be reviewed for possible subtraction from the
sample  spectrum because  of  background  contamination  or  coeluting
peaks.  Data system  library reduction programs can sometimes create
these discrepancies.

                       8260 -  23                         Revision 0
                                                           July  1992

-------
       Computer generated  library search  routines  should  not  use
normalization routines  that would misrepresent the library or unknown
spectra when compared to each other.  Only after visual comparison of
sample  with  the nearest  library  searches  will  the  mass  spectral
interpretation specialist assign a tentative identification.

7.6.2  Quantitative analysis

       7.6.2.1     When  a  compound   has  been   identified,   the
quantitation  of that  compound  will  be  based  on  the  integrated
abundance  from  the   EICP  of  the   primary  characteristic  ion.
Quantitation will  take  place using the internal  standard technique.
The  internal  standard  used  shall  be the  one  nearest  the  retention
time of that of a  given analyte (e.g.  see Table  6).

       7.6.2.2     Calculate  the  concentration  of  each  identified
analyte in the sample as follows:

       Water and Water-Miscible Waste:

                                  (** HI.)
       concentration (M9A) =
                                (AU)(RF)(V0)

where:

       Ax     =     Area  of  characteristic  ion  for  compound  being
                   measured.
       Is     =     Amount  of internal  standard  injected  (ng).
       Ais    =     Area  of   characteristic   ion  for  the  internal
                   standard.
       RF     =     Response   factor  for  compound  being  measured
                   (Section  7.3.6).
       Vo     =     Volume   of   water   purged  (ml),   taking   into
                   consideration any dilutions made.

       Sediment/Soil,  Sludge, and Waste:

       High-concentration:

                                  (*x HI.) (Vt)
       concentration (M9/kg) =
                                (AU)(RF)(V,)(W.J


       Low-concentration:

                                  (A, HU
       concentration (M9Ag) =
                                (AU)(RF)(W.)
                        8260 - 24                        Revision 0
                                                          July  1992

-------
           where:

                  A ,  L, Aic,  RF  =  Same  as  in water  and  water-miscible waste
                   X   S   1 S         ,
                                   above.
                  Vt     =      Volume of  total  extract  (/iL) (use 10,000 juL or a
                              factor of  this when  dilutions  are made).
                  Vj     =      Volume of  extract added  (/^L) for purging.
                  Wj     =      Weight of sample extracted or purged  (g).  The wet
                              weight or  dry weight may be  used, depending upon
                              the specific applications  of the  data.

                  7.6.2.3      Sediment/soil samples are generally reported on a
           dry weight basis,  while  sludges  and wastes  are reported  on  a  wet
           weight  basis.  The  percent dry weight of the sample  (as calculated in
           Section 7.5.3.1.5)  should  be reported  along with the data in either
           instance.

                  7.6.2.4      Where applicable, an estimate of concentration for
           noncalibrated components  in  the sample  should be made.  The formulae
           given above  should be used  with  the   following  modifications:  The
           areas A  and  Ais should be from the total ion chromatograms, and the
           RF for the compound should  be assumed to  be 1.   The concentration
           obtained should  be reported  indicating  (1)  that  the  value  is  an
           estimate and  (2)  which  internal   standard was  used  to determine
           concentration.      Use  the   nearest   internal    standard  free   of
           interferences.
8.0  QUALITY CONTROL

     8.1   Each laboratory  that  uses these methods  is required to  operate a
formal quality control program. The minimum requirements of this program consist
of an initial  demonstration of laboratory  capability  and an ongoing analysis of
spiked  samples  to  evaluate and document quality data.  The  laboratory should
maintain records to document  the quality  of the  data generated.   Ongoing data
quality checks are compared with established performance criteria to determine
if the results of analyses meet the performance characteristics of the method.
When results of sample  spikes  indicate  atypical  method performance,  a quality
control check  sample should be analyzed to  confirm that  the  measurements were
performed in an in-control mode of operation.

     8.2   Before processing any samples, the analyst should demonstrate, through
the  analysis  of a calibration blank,  that  interferences  from  the analytical
system, glassware,  and reagents are under control.  Each time a set of samples
is  extracted  or there  is  a  change  in  reagents,  a  reagent blank  should  be
processed as a  safeguard  against chronic  laboratory  contamination.  The blank
samples  should be carried  through  all   stages  of  sample preparation  and
measurement.

     8.3   The experience  of the analyst performing GC/MS analyses is invaluable
to the success of the methods.  Each day that analysis is performed,  the daily
calibration standard should  be evaluated to determine if the chromatographic
system is operating properly.  Questions that should be asked  are:   Do the peaks
look normal?;  Is the response  obtained comparable to  the response from previous

                                  8260  -  25                         Revision 0
                                                                     July 1992

-------
calibrations?   Careful  examination of the  standard  chromatogram can indicate
whether the column  is still useable, the injector is leaking,  the  injector septum
needs  replacing,  etc.    If  any changes  are made  to  the system  (e.g.  column
changed), recal ibration of the system should take place.

     8.4   Required instrument QC

           8.4.1  The GC/MS system should  be  tuned to meet the BFB specifications
     in Step 7.2.1.

           8.4.2  There should be an  initial calibration of the GC/MS system as
     specified in Section 7.2.

           8.4.3  The  GC/MS  system should meet the  SPCC  criteria specified in
     Section 7.3.3 and the CCC criteria  in Section 7.3.4,  each 12 hours.

     8.5   To establish the ability  to generate acceptable accuracy and precision
on water samples, the analyst should perform the following operations.

           8.5.1  A quality control  (QC) reference sample concentrate is required
     containing each analyte at a concentration of  10 mg/L in methanol .  The QC
     reference sample concentrate  may be  prepared from pure standard materials
     or purchased as certified solutions.   If prepared by the laboratory, the QC
     reference sample concentrate should  be  made using stock standards prepared
     independently from those used  for calibration.

           8.5.2  Prepare a  QC reference sample  to contain  20  /^g/L  of  each
     analyte by adding 200 juL of QC  reference  sample  concentrate to  100 ml of
     water.  For the low level 25  ml a sample,  spike at 5
           8.5.3  Four 5  mL  aliquots  (or  25 ml for low level) of the well-mixed
     QC reference sample  are analyzed according  to the method beginning in Step
     7.4.1.

           8.5.4  Calculate  the  average recovery (R) and the standard deviation
     of the recovery (SJ, for the results.  Ground water background corrections
     should be made before R and RR calculation.

           8.5.5  Tables  7 and 8  provide single laboratory recovery and precision
     data obtained  for the  method analytes from water.   Similar  results from
     dosed water  should  be  expected  by  any  experienced  laboratory.   Compare
     results  obtained in Step  8.5.4 to the single  laboratory recovery  and
     precision  data.   If  the  results  are  not comparable,  review  potential
     problem  areas  and  repeat  the   test.    Results  are  comparable  if  the
     calculated percent  relative  standard deviation  (RSD)  does  not exceed 2.6
     times the single laboratory RSD  or 20%.  whichever is greater and the mean
     recovery lies within the interval  R  ±  3S or R ± 30%, whichever is greater.

           8.5.6  When one or more of the analytes tested fail at least one of
     the acceptance criteria, the analyst  should proceed according to Section
     8.5.6.1 or 8.5.6.2.

                  8.5.6.1     Locate  and  correct the  source of the  problem and
           repeat the test for all analytes beginning with Section  8.5.2.

                                  8260  - 26                         Revision 0
                                                                     July 1992

-------
                  8.5.6.2     Beginning with Section 8.5.2,  repeat the test only
           for those  analytes  that  failed to meet criteria.  Repeated failure,
           however, will confirm  a general problem with the measurement system.
           If this occurs,  locate  and correct  the  source  of the  problem and
           repeat  the test  for all compounds of  interest beginning with Section
           8.5.2.

     8.6   The laboratory should,  on an ongoing basis, analyze a blank and spiked
replicates for each  analytical batch  (up to a maximum  of 20 samples/batch)  to
assess accuracy.  For soil  and waste samples where detectable amounts of organics
are present,  replicate samples may be appropriate in  place of spiked replicates.
For laboratories analyzing  one to  ten samples per month,   at least  one  spiked
sample per month is required.

           8.6.1   The concentration  of   the  spike  in  the sample  should  be
     determined as follows:

                  8.6.1.1     If,  as in compliance monitoring, the concentration
           of a  specific   analyte  in  the  sample is being checked against  a
           regulatory concentration  limit, the spike should be at that limit  or
           1  to 5  times  higher than the  background concentration determined  in
           Step 8.6.2, whichever  concentration would be larger.

                  8.6.1.2     If  the  concentration of a specific analyte  in a
           water sample  is  not being checked against  a specific limit, the spike
           should  be  at  20  ^g/L (or 5 jig/I for  low level) or 1 to 5 times higher
           than the   background  concentration  determined   in  Section  8.6.2,
           whichever  concentration  would be larger.   For  other  matrices,  the
           recommended spiking  concentration  is  10 times the EQL.

           8.6.2   Analyze  one  5 ml  sample aliquot (or 25 ml  for  low level)  to
     determine the background concentration  (B)  of each analyte.   If necessary,
     prepare  a new QC reference  sample concentrate (Section 8.5.1)  appropriate
     for the  background  concentration  in the sample.   Spike  a second  5 ml (or 25
     ml for  low level)  sample aliquot with 10  nL of the  QC reference  sample
     concentrate and  analyze it to determine the  concentration after spiking (A)
     of each  analyte.  Calculate each percent recovery  (p) as 100(A-B)%/T, where
     T is the known true value of the spike.

                  8.6.2.1     Compare  the percent recovery  (R,.) for each analyte
           with QC  acceptance   criteria  established  from  the  analyses  of
           laboratory control  standards   (Section 8.5).  Monitor  all  data from
           dosed samples.   Analyte recoveries should  fall within the established
           control  limits.

                  8.6.2.2      If  recovery is not within limits,  the following
           procedures are  required.

                        8.6.2.2.1   Check  to  be  sure  there  are  no errors  in
                  calculations, matrix spike  solutions  and  internal  standards.
                  Also,  check  instrument  performance.

                        8.6.2.2.2   Recalculate  the  data and/or  reanalyze the
                  extract  if any  of the  above  checks reveal  a problem.

                                   8260  -  27                         Revision 0
                                                                     July 1992

-------
                        8.6.2.2.3    If the checks in 8.6.2.2.1  reveal no errors,
                  the  recovery  problem  encountered with  the  dosed  sample  is
                  judged to be matrix-related,  non system-related.  The result
                  for   that   analyte   in   the   unspiked   sample   is  labeled
                  suspect/matrix to  inform the user that the results are suspect
                  due  to matrix effects.

     8.7   As part of the QC program  for the laboratory, method  accuracy for each
matrix studied should be assessed and records should be maintained.   After the
analysis of five spiked samples _(of the same matrix) as in Step 8.6,  calculate
the  average  percent recovery  (p)  and  the  standard  deviation of the  percent
recovery (sp).  Express the accuracy assessment as a percent recovery interval
from p  - 2s   to  p + 2s .   If p = 90% and s  =  10%,  for example,  the accuracy
interval is  expressed as  70-110%.   Update  the accuracy  assessment for  each
analyte  on  a  regular basis   (e.g.  after  each   five  to ten  new  accuracy
measurements).

     8.8   To determine acceptable accuracy  and precision limits for surrogate
standards the following procedure should be  performed.

           8.8.1  For  each sample analyzed,   calculate  the  percent recovery  of
     each surrogate  in the sample.

           8.8.2  Once a minimum of  thirty samples  of the same matrix have been
     analyzed, calculate the  average percent recovery (p)  and standard deviation
     of the percent  recovery (sp)  for each of the  surrogates.

           8.8.3  For  a given matrix, calculate the upper and lower  control limit
     for method performance for each surrogate standard.  This  should be done as
     follows:

           Upper Control Limit (UCL) =  I + 3s
           Lower Control Limit (LCL) =  p - 3sp

           8.8.4  For  aqueous and soil  matrices,  these laboratory established
     surrogate  control  limits  should,  if  applicable,  be  compared with  the
     control  limits  listed in Table 9.   The  limits given in Table 9 are multi-
     laboratory  performance  based  limits  for   soil  and aqueous  samples,  and
     therefore, the  single-laboratory limits established in  Section 8.8.3 should
     fall within those given in Table 9 for these matrices.

           8.8.5  If recovery is not within  limits,  the following procedures are
     required.

           •       Check to  be  sure  there   are  no   errors   in  calculations,
                  surrogate  solutions  and   internal  standards.   Also,  check
                  instrument performance.

           •       Recalculate the data  and/or reanalyze  the extract  if any  of
                  the  above checks reveal a  problem.

           •       Reextract and reanalyze the sample  if none of the above are a
                  problem or flag  the data  as "estimated concentration".


                                  8260  - 28                        Revision 0
                                                                     July 1992

-------
           8.8.6  At  a minimum, each laboratory should update surrogate recovery
     limits on a matrix-by-matrix basis, annually.

     8.9   It  is  recommended  that  the laboratory  adopt additional  quality
assurance practices for use with this method.  The specific practices that are
most productive depend upon the  needs of  the laboratory and  the nature of the
samples.   Field  duplicates may be  analyzed to  assess  the  precision  of the
environmental measurements. When doubt  exists over the identification of a peak
on the chromatogram,  confirmatory techniques such as gas chromatography with a
dissimilar column  or a different ionization mode using a mass spectrometer should
be used.  Whenever possible,  the laboratory should  analyze  standard reference
materials and participate in relevant performance evaluation  studies.

     8.10  In recognition  of the rapid advances occurring  in chromatography, the
analyst is permitted to modify GC columns, GC conditions, or detectors to improve
the  separations  or  lower  the  cost  of  the measurements.    Each  time  such
modifications to  the  method are made,  the  analyst  is required  to  repeat the
procedure in Section  8.4.


9.0  METHOD PERFORMANCE

     9.1   The  method  detection  limit  (MDL)  is  defined   as  the  minimum
concentration  of  a  substance  that  can  be  measured  and reported with  99%
confidence that the value is above zero.  The MDL actually achieved in a given
analysis will vary depending on instrument sensitivity and matrix effects.

     9.2   This method  has been  tested in  a  single  laboratory  using  spiked
water.  Using a wide-bore capillary column, water was spiked  at concentrations
between 0.5  and 10  M9/L-    Single  laboratory accuracy and precision  data are
presented for the  method analytes in  Table  7.  Calculated MDLs are presented in
Table 1.

     9.3   The  method  was tested using water spiked  at  0.1  to 0.5 ng/L and
analyzed on a cryofocussed narrow-bore column.  The accuracy and precision data
for these compounds are presented in Table 8.  MDL values were also calculated
from these data and are presented in Table 2.


10.0  REFERENCES

1.   Methods for  the  Determination of  Organic Compounds  in  Finished  Drinking
     Water  and  Raw  Source Water Method 524.2;  U.S.  Environmental  Protection
     Agency.  Office   of   Research  Development.   Environmental  Monitoring  and
     Support Laboratory: Cincinnati,  OH 1986.

2.   U.S.  EPA  Contract   Laboratory  Program, Statement  of  Work  for  Organic
     Analysis, July 1985,  Revision.

3.   Bellar, T.A.; Lichtenberg, J.J.  vL Amer. Water Works Assoc. 1974,  66(12),
     739-744.

4.   Bellar,  T.A.;  Lichtenberg,  J.J.   "Semi-Automated  Headspace Analysis  of
     Drinking  Waters  and Industrial Waters for  Purgeable  Volatile  Organic

                                   8260  -  29                        Revision 0
                                                                     July 1992

-------
     Compounds";  in Van Hall, Ed.; Measurement  of Organic  Pollutants In Water
     and Wastewater.  ASTM STP 686,  pp 108-129, 1979.

5.   Budde,  W.L.; Eichelberger,  J.W.  "Performance Tests  for the Evaluation of
     Computerized   Gas   Chromatography/Mass   Spectrometry   Equipment   and
     Laboratories";  U.S.   Environmental   Protection  Agency.   Environmental
     Monitoring and Support Laboratory.  Cincinnati, OH 45268, April 1980; EPA-
     600/4-79-020.

6.   Eichelberger,  J.W.;  Harris,  L.E.;   Budde,  W.L.  "Reference  Compound  to
     Calibrate Ion  Abundance Measurement in Gas Chromatography-Mass Spectrometry
     Systems"; Analytical Chemistry 1975,  47, 995-1000.

7.   Olynyk, P.; Budde,  W.L.;  Eichelberger,  J.W.  "Method  Detection  Limit  for
     Methods 624 and  625"; Unpublished report, October 1980.

8.   Non   Cryogenic   Temperatures   Program   and    Chromatogram,    Private
     Communications;  Myron Stephenson  and Frank Allen, EPA Region  IV Laboratory,
     Athens, GA.
                                   8260 - 30                        Revision 0
                                                                     July  1992

-------
                            TABLE  1.
CHROMATOGRAPHIC RETENTION TIMES AND METHOD DETECTION LIMITS (MDL)
  FOR VOLATILE ORGANIC COMPOUNDS ON WIDE BORE CAPILLARY COLUMNS
ANALYTE
                                      RETENTION TIME
                                         (minutes)
     MDLd

Di chl orodi f 1 uoromethane
Chl oromethane
Vinyl Chloride
Bromomethane
Chloroethane
Tri chl orof 1 uoromethane
1,1-Dichloroethene
Methyl ene chloride
trans -1,2-Di chl oroethene
1,1-Dichloroethane
2,2-Dichloropropane
cis- 1,2-Di chl oroethene
Chloroform
Bromochl oromethane
1,1, 1 -Trichl oroethane
Carbon tetrachloride
1,1-Dichloropropene
Benzene
1,2-Di chl oroethane
Trichloroethene
1,2-Dichloropropane
Bromodi chl oromethane
Dibromomethane
trans- 1,3-Dichloropropene
Toluene
cis- 1,3-Dichloropropene
1,1,2-Trichloroethane
Tetrachl oroethene
1,3-Dichloropropane
Di bromochl oromethane
1,2-Dibromoethane
1-Chlorohexane
Chlorobenzene
1,1,1 , 2-Tetrachl oroethane
Ethyl benzene
p-Xylene
m-Xylene
o-Xylene
Styrene
Bromoform
Isopropyl benzene
1,1,2, 2-Tetrachl oroethane
Column la
1.55
1.63
1.71
2.01
2.09
2.27
2.89
3.60
3.98
4.85
6.01
6.19
6.40
6.74
7.27
7.61
7.68
8.23
8.40
9.59
10.09
10.59
10.65
--
12.43
--
13.41
13.74
14.04
14.39
14.73
15.46
15.76
15.94
15.99
16.12
16.17
17.11
17.31
17.93
18.06
18.72
Column 2°
0.70
0.73
0.79
0.96
1.02
1.19
1.57
2.06
2.36
2.93
3.80
3.90
4.80
4.38
4.84
5.26
5.29
5.67
5.83
7.27
7.66
8.49
7.93
--
10.00
--
11.05
11.15
11.31
11.85
11.83
13.29
13.01
13.33
13.39
13.69
13.68
14.52
14.60
14.88
15.46
16.35
Column 2"
3.13
3.40
3.93
4.80
--
6.20
7.83
9.27
9.90
10.80
11.87
11.93
12.60
12.37
12.83
13.17
13.10
13.50
13.63
14.80
15.20
15.80
15.43
16.70
17.40
17.90
18.30
18.60
18.70
19.20
19.40
--
20.67
20.87
21.00
21.30
21.37
22.27
22.40
22.77
23.30
24.07

0.10
0.13
0.17
0.11
0.10
0.08
0.12
0.03
0.06
0.04
0.35
0.12
0.03
0.04
0.08
0.21
0.10
0.04
0.06
0.19
0.04
0.08
0.24
--
0.11
--
0.10
0.14
0.04
0.05
0.06
0.05
0.04
0.05
0.06
0.13
0.05
0.11
0.04
0.12
0.15
0.04
                            8260 - 31
Revision 0
 July 1992

-------
                                   TABLE 1.
                                  (Continued)
ANALYTE
         RETENTION TIME
           (minutes)
                                    Column T
            Column 2    Column 2
                                                                     /c
INTERNAL STANDARDS/SURROGATES

4-Bromofluorobenzene
18.63
15.71
23.63
                         MDL"
Bromobenzene
1,2,3-Trichloropropane
n-Propyl benzene
2-Chlorotoluene
1, 3 ,5-Tri methyl benzene
4-Chlorotoluene
tert-Butyl benzene
1 , 2 , 4-Trimethyl benzene
sec-Butyl benzene
p- I sopropyl toluene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
n-Butyl benzene
1,2-Dichlorobenzene
l,2-Dibromo-3-chloropropane
1,2,4-Trichlorobenzene
Hexachlorobutadiene
Naphthalene
1,2,3-Trichlorobenzene
18.95
19.02
19.06
19.34
19.47
19.50
20.28
20.34
20.79
21.20
21.22
21.55
22.22
22.52
24.53
26.55
26.99
27.17
27.78
15.86
16.23
16.41
16.42
16.90
16.72
17.57
17.70
18.09
18.52
18.14
18.39
19.49
19.17
21.08
23.08
23.68
23.52
24.18
24.00
24.13
24.33
24.53
24.83
24.77
26.60
31.50
26.13
26.50
26.37
26.60
27.32
27.43
--
31.50
32.07
32.20
32.97
0.03
0.32
0.04
0.04
0.05
0.06
0.14
0.13
0.13
0.12
0.12
0.03
0.11
0.03
0.26
0.04
0.11
0.04
0.03
a  Column 1 - 60 meter x 0.75 mm ID VOCOL capillary.  Hold at 10°C for 5  minutes,
   then program  to  160°C at 6°/nnn.

b  Column 2 - 30 meter x 0.53 mm  ID DB-624 wide-bore capillary  using cryogenic
   oven.  Hold at 10°C for 5 minutes, then program to 160°C at 6°/min.

c  Column 2' - 30 meter x 0.53 mm  ID DB-624 wide-bore capillary,  cooling GC oven
   to  ambient  temperatures.  Hold  at 10°C for 6 minutes,  program to  70°C  at
   10°/min, program to 120°C at  5°/min,  then  program to  180°C  at  8°/min.

d  MDL based on  a 25  mL sample volume.
                                   8260 - 32
                                Revision 0
                                 July 1992

-------
                            TABLE 2.
CHROMATOGRAPHIC RETENTION TIMES AND METHOD DETECTION LIMITS (MDL)
 FOR VOLATILE ORGANIC COMPOUNDS ON NARROW BORE CAPILLARY COLUMNS
ANALYTE
Di chl orodi f 1 uoromethane
Chloromethane
Vinyl chloride
Bromomethane
Chloroethane
Tri chl orof 1 uoromethane
1,1-Dichloroethene
Methylene chloride
trans- 1,2-Di chl oroethene
1,1-Dichloroethane
cis- 1,2-Di chl oroethene
2,2-Dichloropropane
Chloroform
Bromochl oromethane
1,1,1 -Tri chl oroethane
1,2-Dichloroethane
1,1-Dichloropropene
Carbon tetrachloride
Benzene
1,2-Dichloropropane
Trichloroethene
Dibromomethane
Bromod i chl oromethane
Toluene
1,1,2-Trichloroethane
1,3-Dichloropropane
Di bromochl oromethane
Tetrachl oroethene
1,2-Di bromoethane
Chlorobenzene
1,1,1 , 2-Tetrachl oroethane
Ethyl benzene
p-Xylene
m-Xylene
Bromoform
o-Xylene
Styrene
1,1,2 , 2-Tetrachl oroethane
1,2, 3 -Tri chl oropropane
I sopropyl benzene
RETENTION TIME
(minutes)
Column 3
0.88
0.97
1.04
1.29
1.45
1.77
2.33
2.66
3.54
4.03
5.07
5.31
5.55
5.63
6.76
7.00
7.16
7.41
7.41
8.94
9.02
9.09
9.34
11.51
11.99
12.48
12.80
13.20
13.60
14.33
14.73
14.73
15.30
15.30
15.70
15.78
15.78
15.78
16.26
16.42
MDLb
(M9/L)
0.11
0.05
0.04
0.06
0.02
0.07
0.05
0.09
0.03
0.03
0.06
0.08
0.04
0.09
0.04
0.02
0.12
0.02
0.03
0.02
0.02
0.01
0.03
0.08
0.08
0.08
0.07
0.05
0.10
0.03
0.07
0.03
0.06
0.03
0.20
0.06
0.27
0.20
0.09
0.10
                            8260  -  33
Revision 0
 July 1992

-------


ANALYTE


Bromobenzene
2-Chlorotoluene
n-Propyl benzene
4-Chlorotoluene
1 , 3 , 5-Trimethyl benzene
tert-Butyl benzene
1 , 2 , 4-Trimethyl benzene
sec-Butyl benzene
1,3-Dichlorobenzene
p- I sopropyl tol uene
1,4-Dichlorobenzene
1,2-Dichlorobenzene
n-Butyl benzene
1 , 2-Di bromo-3-chl oropropane
1 , 2 , 4-Tri chl orobenzene
Naphthalene
Hexachl orobutadi ene
1 , 2 , 3-Tri chl orobenzene
TABLE 2.
(Continued)
RETENTION TIME
(minutes)
/* 1 M. *5«
Column 3
16.42
16.74
16.82
16.82
16.99
17.31
17.31
17.47
17.47
17.63
17.63
17.79
17.95
18.03
18.84
19.07
19.24
19.24


MDLb
(M9/L)

0.11
0.08
0.10
0.06
0.06
0.33
0.09
0.12
0.05
0.26
0.04
0.05
0.10
0.50
0.20
0.10
0.10
0.14
a  Column 3 - 30 meter x 0.32 mm ID DB-5 capillary with 1

b  MDL based on a 25 ml sample volume.
                           film  thickness.
8260 - 34
                                                                    Revision 0
                                                                     July 1992

-------
                                 TABLE 3.
           ESTIMATED QUANTITATION LIMITS FOR VOLATILE ANALYTES3
                                            Estimated
                                           Quantitation
                                              Limits
                                  Ground waterLow Soil/Sediment0
    Volume of water purged        5 mL     25 mL

    All  analytes in Table 1       5        1
   Estimated Quantitation Limit (EQL)  - The lowest concentration that can be
   reliably  achieved within specified limits of precision and accuracy during
   routine   laboratory  operating  conditions. The  EQL  is  generally  5  to 10
   times the MDL.  However, it may  be nominally  chosen within these guidelines
   to  simplify  data    reporting.   For   many  analytes   the   EQL  analyte
   concentration  is  selected  for  the   lowest  non-zero   standard  in  the
   calibration  curve.  Sample EQLs  are highly   matrix-dependent. The  EQLs
   listed herein are  provided for  guidance  and  may  not always be achievable.
   See the  following information  for  further guidance  on matrix-dependent
   EQLs.

   EQLs listed  for soil/sediment  are  based on wet weight.  Normally  data is
   reported on a dry weight  basis;  therefore,  EQLs will be higher,  based on
   the percent dry weight in each  sample.
             Other Matrices                      Factor0


             Water miscible liquid waste             50
             High-concentration soil and sludge     125
             Non-water miscible waste               500


CEQL =  [EQL for  low  soil  sediment  (Table  3)]  X  [Factor].  For non-aqueous
        samples, the factor is on a wet-weight basis.
                                 8260  -  35                         Revision 0
                                                                   July 1992

-------
                             TABLE 4.
    BFB MASS - INTENSITY SPECIFICATIONS (4-BROMOFLUOROBENZENE)


Mass              Intensity Required (relative abundance)
 50               15 to 40% of mass 95
 75               30 to 60% of mass 95
 95               base peak,  100% relative  abundance
 96               5 to 9% of mass 95
173               less than 2% of mass 174
174               greater than 50% of mass  95
175               5 to 9% of mass 174
176               greater than 95% but less than  101% of mass  174
177               5 to 9% of mass 176
                            8260 - 36                         Revision  0
                                                               July  1992

-------
                             TABLE 5.
   CHARACTERISTIC MASSES  (M/Z)  FOR  PURGEABLE  ORGANIC  COMPOUNDS'
                                Primary           Secondary
                              Characteristic    Characteristic
Analyte                           Ion               Ion(s)
Benzene
Bromobenzene
Bromochl oromethane
Bromodichloromethane
Bromoform
Bromomethane
n-Butyl benzene
sec-Butyl benzene
tert-Butyl benzene
Carbon tetrachloride
Chlorobenzene
Chloroethane
Chloroform
Chi oromethane
2-Chlorotoluene
4-Chlorotoluene
l,2-Dibromo-3-chloropropane
Di bromochl oromethane
1,2-Dibromoethane
Dibromomethane
1 , 2-Di chl orobenzene
1,3-Dichlorobenzene
1 ,4-Di chlorobenzene
Dichlorodi fluoromethane
1,1-Dichloroethane
1,2-Dichloroethane
1,1-Dichloroethene
cis-l,2-D'ch1oroethene
trans -1 , 2- Oichl oroethene
1,2-Dichloropropane
1,3-Dichloropropane
2,2-Dichloropropane
1,1-Dichloropropene
Ethyl benzene
Hexachlorobutadiene
Isopropyl benzene
p-Isopropyl toluene
Methylene chloride
Naphthalene
n-Propyl benzene
Styrene
1,1,1 , 2-Tetrachl oroethane


78
156
128
83
173
94
91
105
119
117
112
64
83
50
91
91
75
129
107
93
146
146
146
85
63
62
96
96
96
63
76
77
75
91
225
105
119
84
128
91
104
131
8260 - 37

.
77, 158
49, 130
85, 127
175, 254
96
92, 134
134
91, 134
119
77, 114
66
85
52
126
126
155, 157
127
109, 188
95, 174
111, 148
111, 148
111, 148
87
65, 83
98
61, 63
61, 98
61, 98
112
78
97
110, 77
106
223, 227
120
134, 91
86, 49
-
120
78
133, 119
Revision 0
July 1992

-------
                                   TABLE 5.
                                  (Continued)
                                      Primary           Secondary
                                    Characteristic    Characteristic
      Analyte                           Ion               Ion(s)
1,1,2 , 2-Tetrachl oroethane
Tetrachloroethene
Toluene
1,2,3-Trichlorobenzene
1,2,4-Trichlorobenzene
1,1,1 -Trichl oroethane
1 , 1 , 2-Tri chl oroethane
Trichloroethene
Trichl orofluoromethane
1 ,2,3-Trichloropropane
1 , 2 , 4-Tri methyl benzene
1,3, 5-Trimethyl benzene
Vinyl chloride
o-Xylene
m-Xylene
p-Xylene
83
166
92
180
180
97
83
95
101
75
105
105
62
106
106
106
131,
168,
91
182,
182,
99,
97,
130,
103
77
120
120
64
91
91
91
85
129

145
145
61
85
132








INTERNAL STANDARDS/SURROGATES

      4-Bromofluorobenzene              95               174,  176
      Dibromofl loromethane             113
      To1uene-d8                         98
      Pentafluorobenzene               168
      1,4-Difluorobenzene              114
      Ch1orobenzene-d5                  117
      l,4-Dichiorobenzene-d4            152
                                  8260  -  38                         Revision 0
                                                                     July 1992

-------
                                   TABLE 6.
            VOLATILE  INTERNAL  STANDARDS  WITH CORRESPONDING ANALYTES
                           ASSIGNED  FOR  QUANTITATION
Pentafluorobenzene

Acetone
Acrolein
Acrylonitrile
Bromochloromethane
Bromomethane
2-Butanone
Carbon dlsulfide
Chloroethane
Chloroform
Chloromethane
Di chlorodi f1uoromethane
1,1-Dichloroethane
1,1-Dichloroethene
ci s-1,2-Dichloroethene
trans-1,2-Dichloroethene
2,2-Dichloropropane
lodomethane
Methylene chloride
1,1,1-Tri chloroethane
Tri chlorof1uoromethane
Vinyl acetate
Vinyl chloride

Chlorobenzene-dr

Bromoform
Chlorodibromomethane
Chlorobenzene
1,3-Dichloropropane
Ethyl benzene
2-Hexanone
Styrene
1,1,1,2-Tetrachloroethane
Tetrachloroethene
Xylene
1,4-Difluorobenzene

Benzene
Bromodichloromethane
Bromofluorobenzene (surrogate)
Carbon tetrachloride
2-Chloroethyl vinyl ether
1,2-Dibromoethane
Dibromomethane
1,2-Dichloroethane
1,2-Dichloroethane-d4 (surrogate)
1,2-Dichloropropane
1,1-Dichloropropene
cis-l,3-Dichloropropene
trans-l,3-Dichloropropene
4-Methyl-2-pentanone
Toluene
Toluene-d8 (surrogate)
1,1,2-Trichloroethane
Trichloroethene

1.4-Dichlorobenzene-d4

Bromobenzene
n-8utylbenzene
sec-Butylbenzene
tert-Butylbenzene
2-Chlorotoluene
4-Chlorotoluene
1,2-Di bromo-3-chloropropane
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
Hexachlorobutadiene
Isopropyl benzene
p-Isopropyltoluene
Naphthalene
n-Propylbenzene
1,1,2,2-Tetrachloroethane
1,2,3-Trichlorobenzene
1,2,4-Trichlorobenzene
1,2,3-Tri chloropropane
1,2,4-Trimethyl benzene
1,3,5-Trimethyl benzene
                                   8260 - 39
                                Revision 0
                                 July 1992

-------
                         TABLE 7.
SINGLE LABORATORY ACCURACY AND PRECISION DATA FOR VOLATILE
    ORGANIC COMPOUNDS  IN WATER DETERMINED WITH A WIDE
                  BORE CAPILLARY COLUMN
Analyte
Benzene
Bromobenzene
Bromochl oromethane
Bromodi chl oromethane
Bromoform
Bromomethane
n-Butyl benzene
sec-Butyl benzene
tert -Butyl benzene
Carbon tetrachloride
Chlorobenzene
Chloroethane
Chloroform
Chl oromethane
2-Chlorotoluene
4-Chlorotoluene
1 , 2-Dibromo-3-Ch1 oropropane
Di bromochl oromethane
1,2-Dibromoethane
Dibromomethane
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
Di chl orod i f 1 uoromethane
l,l-Dichloroben?ene
1,2-Dichlorobenzene
1,1-Dichloroethene
cis-l,2-Dichloroethene
trans- 1,2-Di chl oroethene
1,2-Dichl oropropane
1,3-Dichloropropane
2,2-Dichloropropane
1,1-Dichloropropene
Ethyl benzene
Hexachl orobutadi ene
I sopropyl benzene
p-Isopropyl toluene
Methylene chloride
Naphthalene
n-Propyl benzene


Cone. Number
Range, of Recovery,8
jig/L Samples %
0.1
0.1
0.5
0.1
0.5
0.5
0.5
0.5
0.5
0.5
0.1
0.5
0.5
0.5
0.1
0.1
0.5
0.1
0.5
0.5
0.1
0.5
0.2
0.5
0.5
0.1
0.1
0.5
0.1
0.1
0.1
0.5
0.5
0.1
0.5
0.5
0.1
0.1
0.1
0.1


- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 20
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
-100
- 10
8260

31
30
24
30
18
18
18
16
18
24
31
24
24
23
31
31
24
31
24
24
31
24
31
18
24
31
34
18
30
30
3T
12
18
31
18
16
23
30
31
31
- 40

97
100
90
95
101
95
100
100
102
84
98
89
90
93
90
99
83
92
102
100
93
99
103
90
96
95
94
101
93
97
96
86
98
99
100
101
99
95
104
100


Standard Percent
Deviation Rel . Std.
of Recovery6 Dev.
6.5
5.5
5.7
5.7
6.4
7.8
7.6
7.6
7.4
7.4
5.8
8.0
5.5
8.3
5.6
8.2
16.6
6.5
4.0
5.5
5.8
6.8
6.6
6.9
5.1
5.1
6.3
6.7
5.2
5.9
5.7
14.6
8.7
8.4
6.8
7.7
6.7
5.0
8.6
5.8


5.7
5.5
6.4
6.1
6.3
8.2
7.6
7.6
7.3
8.8
5.9
9.0
6.1
8.9
6.2
8.3
19.9
7.0
3.9
5.6
6.2
6.9
6.4
7.7
5.3
5.4
6.7
6.7
5.6
6.1
6.0
16.9
8.9
8.6
6.8
7.6
6.7
5.3
8.2
5.8
Revision 0
July 1992

-------
                                   TABLE 7.
                                  (Continued)
Analyte
Cone.   Number
Range,    of
/ig/L    Samples
Recovery,'
   %
  Standard
 Deviation
of Recovery6
  Percent
Rel. Std.
     Dev.
Styrene
1,1,1 , 2-Tetrachl oroethane
1,1,2, 2-Tetrachl oroethane
Tetrachloroethene
Toluene
1,2,3-Trichlorobenzene
1,2,4-Trichlorobenzene
1,1,1-Trichl oroethane
1 , 1 , 2-Trichloroethane
Trichloroethene
Trichlorofluoromethane
1,2,3-Trichloropropane
1 , 2, 4-Trimethyl benzene
1,3, 5-Trimethyl benzene
Vinyl chloride
o-Xylene
m-Xylene
p-Xylene
0.1
0.5
0.1
0.5
0.5
0.5
0.5
0.5
0.5
0.5
0.5
0.5
0.5
0.5
0.5
0.1
0.1
0.5
-100
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 31
- 10
- 10
39
24
30
24
18
18
18
18
18
24
24
16
18
23
18
18
31
18
102
90
91
89
102
109
108
98
104.
90
89
108
99
92
98
103
97
104
7.3
6.1
5.7
6.0
8.1
9.4
9.0
7.9
7.6
6.5
7.2
15.6
8.0
6.8
6.5
7.4
6.3
8.0
7.2
6.8
6.3
6.8
8.0
8.6
8.3
8.1
7.3
7.3
8.1
14.4
8.1
7.4
6.7
7.2
6.5
7.7
9  Recoveries  were  calculated   using   internal   standard  method.  Internal
   standard was fluorobenzene.

b  Standard   deviation   was   calculated   by   pooling   data   form   three
   concentrations.
                                   8260 - 41
                                        Revision 0
                                         July 1992

-------
                    TABLE 8.
SINGLE LABORATORY ACCURACY AND PRECISION DATA FOR
 VOLATILE ORGANIC COMPOUNDS IN WATER DETERMINED
       WITH A NARROW BORE CAPILLARY COLUMN
Analyte
Benzene
Bromobenzene
Bromochloromethane
Bromodichloromethane
Bromoform
Bromomethane
n -Butyl benzene
sec-Butyl benzene
tert-Butyl benzene
Carbon tetrachloride
Chlorobenzene
Chloroethane
Chloroform
Chloromethane
2-Chlorotoluene
4-Chlorotoluene
1 , 2-Di bromo-3-chl oropropane
Di bromochl oromethane
1,2-Dibromoethane
Dibromomethane
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
Di chl orodi f 1 uoromethane
1,1-Dichloroethane
1,2-Dichloroethane
1,1-Dichloroethene
cis-1, 2-Di chl oroethene
trans - 1 , 2 -Di chl oroethene
1,2-Dichloropropane
1,3-Dichloropropane
2,2-Dichloropropane
1,1-Dichloropropene
Ethyl benzene
Hexachl orobutadi ene
Isopropyl benzene
p- I sopropyl toluene
Methyl ene chloride
Naphthalene
n-Propylbenzene


Cone.
M9/L
0.1
0.5
0.5
0.1
0.5
0.5
0.5
0.5
0.5
0.1
0.1
0.1
0.1
0,5
0.5
0.5
0.5
0.1
0.5
0.5
0.1
0.1
0.1
0.1
0.5
0.1
0.1
0.1
0.1
0.5
0.5
0.5
0.5
0.5
0.5
0.5
0.5
0.5
0.5
0.5


Number
of
Samples
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
8260 - 42

Recovery,8
%
99
97
97
100
101
99
94
110
110
108
91
100
105
101
99
96
92
99
97
93
97
101
106
99
98
100
95
100
98
96
99
99
102
99
100
102
113
97
98
99


Standard
Deviation
of Recovery
6.2
7.4
5.8
4.6
5.4
7.1
6.0
7.1
2.5
6.8
5.8
5.8
3.2
4.7
4.6
7.0
10.0
5.6
5.6
5.6
3.5
6.0
6.5
8.8
6.2
6.3
9.0
3.7
7.2
6.0
5.8
4.9
7.4
5.2
6.7
6.4
13.0
13.0
7.2
6.6


Percent
Rel. Std.
Dev.
6.3
7.6
6.0
4.6
5.3
7.2
6.4
6.5
2.3
6.3
6.4
5.8
3.0
4.7
4.6
7.3
10.9
5.7
5.8
6.0
3.6
5.9
6.1
8.9
6.3
6.3
9.5
3.7
7.3
6.3
5.9
4.9
7.3
5.3
6.7
6.3
11.5
13.4
7.3
6.7
Revision 0
July 1992

-------
TABLE 8.
(Continued)


Analyte
Styrene
1,1,1 , 2-Tetrachl oroethane
1,1,2 , 2-Tetrachl oroethane
Tetrachloroethene
Toluene
1,2,3-Trichlorobenzene
1,2,4-Trichlorobenzene
1,1,1 -Tri chl oroethane
1 , 1 , 2-Trichl oroethane
Trichloroethene
Tr i chl orof 1 uoromethane
1,2,3-Trichloropropane
1 , 2 , 4-Trimethyl benzene
1,3 , 5-Trimethyl benzene
Vinyl chloride
o-Xylene
m-Xylene
p-Xylene

Cone.
M9/L
0.5
0.5
0.5
0.1
0.5
0.5
0.5
0.5
0.5
0.1
0.1
0.5
0.5
0.5
0.1
0.5
0.5
0.5
Number
of
Samples
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7

Recovery,8
%
96
100
100
96
100
102
91
100
102
104
97
96
96
101
104
106
106
97
Standard
Deviation
of Recovery
19.0
4.7
12.0
5.0
5.9
8.9
16.0
4.0
4.9
2.0
4.6
6.5
6.5
4.2
0.2
7.5
4.6
6.1
Percent
Rel. Std.
Dev.
19.8
4.7
12.0
5.2
5.9
8.7
17.6
4.0
4.8
1.9
4.7
6.8
6.8
4.2
0.2
7.1
4.3
6.3
Recoveries  were  calculated
standard was fluorobenzene.
using  internal  standard  method.   Internal
                                8260 - 43
                                   Revision 0
                                    July 1992

-------
                                   TABLE 9.
      SURROGATE  SPIKE  RECOVERY  LIMITS  FOR  WATER AND SOIL/SEDIMENT SAMPLES
Surrogate Compound
4-Bromofl uorobenzene8
Di bromof 1 uoromethane8
Toluene-d8a
Low/High
Water
86-115
86-118
88-110
Low/High
Soil/Sediment
74-121
80-120
81-117
8  Single laboratory data for guidance only.
                                   TABLE  10.

                 QUANTITY OF EXTRACT REQUIRED FOR ANALYSIS OF
                          HIGH-CONCENTRATION SAMPLES
Approximate                                     Volume of
Concentration Range                             Extract8
   500 -  10,000 /ig/kg                          100
 1,000 -  20,000 pig/kg                           50
 5,000 - 100,000 /ig/kg                           10
25,000 - 500,000 /ig/kg                          100 /xL of 1/50 dilution6


Calculate appropriate dilution factor for concentrations exceeding this table.

a     The volume of solvent added to 5 mL of water being purged should be kept
      constant.  Therefore, add to the 5 mL syringe whatever volume of solvent
      is necessary to maintain a volume of 100 nl added to the syringe.

b     Dilute  an aliquot  of  the  solvent  extract  and  then  take  100 /uL  for
      analysis.
                                   8260  -  44                        Revision 0
                                                                     July 1992

-------
                         FIGURE 1.
                      PURGING  DEVICE
                        CUT <* « OA
                         FIGURE 2.
TRAP PACKING AND CONSTRUCTION TO INCLUDE DESORB CAPABILITY
              0«0<1MO OtTM.
                         8260  -  45
Revision 0
 July 1992

-------
                      FIGURE 3.
SCHEMATIC  OF PURGE-ANO-TRAP  DEVICE  - PURGE MODE
  CAWWOAS
  aow CONTWOL
  •UUCOAI
  nowcoMTKx
  mMOLfCULAft
  MVf W.TCT
                                     NOTt
                                     »U. UHf» KTV«f W m»»
                                     ANO oc SHOULD M MCATV
                       FIGURE  4.
SCHEMATIC OF PURGE-AND-TRAP DEVICE -  DESORB  MODE
   CAANOOMI
   FLOWCONTMX
                               COLUMN OVtN

                                      conrmMATpm COLUMN

                                      TO UtTCCTDft
                                     *u uwts inxoH TIM*
                                     AMD OC SHOULD M HtATB)
                                     TOVC
                       8260  -  46
Revision 0
 July 1992

-------
  00
  ro
  en
  o
                                                 JL.
          COLUMN!  &0 METER » O.73 NN I, O.  VQCOl. Cf*PIt,UU*V


                    10 c FOR a HIN.,  TMSH  4 /KIN TO t&o C
                                                                                         2000
                                                                                        -.  i  .
24OO
                                                  NCTCNIIOM IMMf. MM.
o o>
c <
U? 3
\£>
r\J O

-------
  00
  ro
  en
  o
  CO
         PHOOHAHi  IO C  FOR 9 HIM.,

                  1MEN 6  /NIN  TO  I6O C
                  V«-| f  r  i  I  •  i i
  30
C-. (B
C <
                                                       AtlfMIIOM IMM. MM.
                                                                                                                                O
                                                                                                                                £75
  tP
O C
-n po
                                                                                                                                o
                                                                                                                                33
                                                                                                                                CT>
^- O
VO 3

-------
             R1C                                  DATA:  -JO»'S0423c.7 1843    SCANS   125 TO  900
             04-23--87   y:2s:00                   CALI:  46US042y8? 13
             SAt-lFlE: 40t'OAST00423dr, 5UL.-5ML
             CGHGS.: F4000,40-160X8,12,F4,30MLPURGE,TENSILGEL.Dee24.SWEEP35,10FS1
             RANGE:  G    1.1200  LABEL: H  0, 4.0   GUAM:  A  0, 1.0 J  6  BASE:  U  20,   3
     lOO.O-i
oo
ro
cri
o
.p.
VD
      R1C
 73

' <

 (/>

' O
i 3
                                a
                     1	
                     "flfl
                                   3«u
~~l	
 400
 6:40
~]	
 500
 9:20
—I	
  600
 10:00
—I	
  700
 11:40
—T""
  800
 13:20
                                                                                                                          264132.
                                                                                                                                   oo
                                                                                                                                   o
                                                                                                                                   33
                                                                                                                                   O
                                                                                                                                    O
                                                                                                                                    G1
                                                                                                                                    o
	1
  »00>  SCHl/
 15:00 TIME

-------
            1
 00
 ro
 cr>
 O
 en
 O
C_ (D
C <
                                            U
 $
M
                             si      *

                        I    11      1
                                                  5
                                                                                                        I. l.l-*OMlOMOrimiOlt

                                                                                                        1.
                                                               • CMUMOT

                                                               » «.«.*-
                                                                                                        ti. n
                                                                           M(Mff1»)
                                                              II IfMOttAftOOMV

                                                              IJ IJ
                                                                                                        «*  VOUKMI

                                                              M. nNlUHMICMt

                                                              ft. t.»-V«UM(

                                                              1>.
                                                              •»

                                                              M

                                                                                                                                              '00
                                                                                                                                             X
                                                                                                                                             —1
                                                                                                                                             cr
I)
                              U

                             JL.
~ O
VO 3
ro O

-------
                 FIGURE 9.
            LOW SOILS IMPINGER
 PURGE INLET FITTING
                                      SEPTUM
J  i |mm 0 0 GLASS
             40ml VIAL
                 8260  - 51
Revision  0
 July 1992

-------
                                      METHOD 8250
VOLATILE  ORGANIC COMPOUNDS  BY GAS CHROMATOGRAPHY/MASS  SPECTROMETRY  (GC/MS):
                             CAPILLARY COLUMN  TECHNIQUE
                       Purge-and-trap
                                  7 1
                                 S.l.ct
                              pr ocedure for
                               in I reducing
                               •ample into
                                CC/MS
                               732 Tuna
                              CC/MS lyjtem
                                • Uh BF8
                               732/733
                               A»»embl •
                             purge•and•trap
                             device   Conned
                              device to CC
                              734 Prepare
                               c-al i br* t 2 on
                               •tandardl
                              736 Perform
                             purge•and•trap
                                ana 1ysi*
    7 3 7
Calculate RFi
for  5 SPCCi
                                                       7 3 8
                                                     Calculate
                                                    *RSD of Rr
                                                     for CCC»
7 4 Perform
daily
calibration


                                      8260 -  52
                          Revision 0
                           July 1992

-------
                                    METHOD  8260
                                    (Continued)
                 Low concentratio
                 soil /seaiment
 7 5 31.3  Prepare
aqueou*  solution of
   surrogate and
internal  standard*
                   water and wa ter
                   mi scible 1 iquids
751/752 Screen
sample  us ing Method
   3810 or  382C
      (Dilute
  water-miscible
liquid* at  least SO
      fold  ]
                             7  S  1  3 Add
                          internal  standard
                            and surroga te
                         spitting  solution*
                          7 5 1  10/7 £ 1 11
                               Par form
                           purge -and- 1 rap
                             pr oodur •
                                                     7  S3  22 Add
                                                   sol vent,  internal
                                                     standard and
                                                   surrogate spiking
                                                   solution*  Shake
                                                    7532  Store
                                                  portion  of extract
                                                   for  re-analy*i*
                                                    Prepare method
                                                        blank
                                                                        7 5 1  12  Desorb trap
                                                                        on to column  Ana 1y xe
                                                                               samole
                                                                        chromatographically
 7611  Identify
    analytes by
   coaparing the
sample  and standard
   mass spectra.
                                                                                I
                                                 7622 Calculate
                                                 the concentration
                                                of  each identified
                                     8260  -  53
                                                           Revision  0
                                                             July  1992

-------
00
to

-------
                                 METHOD 8260A

 VOLATILE ORGANIC  COMPOUNDS  BY GAS CHROMATOGRAPHY/MASS  SPECTROMETRY  (GC/MS):
                          CAPILLARY COLUMN TECHNIQUE
1.0   SCOPE AND APPLICATION

      1.1   Method 8260  is  used  to determine volatile organic  compounds  in a
variety of solid waste matrices.  This method  is applicable to nearly all types
of samples,  regardless of water content, including ground water, aqueous sludges,
caustic  liquors,  acid liquors,  waste  solvents,  oily wastes
fibrous  wastes,  polymeric  emulsions,   filter  cakes,  spent
catalysts, soils, and sediments.   The following compounds can
this method:
                          mousses,  tars,
                          carbons,   spent
                         be  determined  by
Analyte
CAS No.1
  Appropriate Technique
                  Direct
Purge-and-Trap    Injection
Acetone
Acetonitrile
Acrolein (Propenal)
Acrylonitrile
Allyl alcohol
Allyl chloride
Benzene
Benzyl chloride
Bromoacetone
Bromochloromethane (I.S.)
Bromodichloromethane
4-Bromofluorobenzene (surr.)
Bromoform
Bromomethane
n-Butanol
2-Butanone (MEK)
Carbon disulfide
Carbon tetrachloride
Chloral hydrate
Chlorobenzene
2-Chloro-l,3-butadiene
Chi orodi bromomethane
Chloroethane
2-Chloroethanol
bis-(2-Chloroethyl) sulfide
2-Chloroethyl vinyl ether
Chloroform
Chloromethane
Chloroprene
67-64-1
75-05-8
107-02-8
107-13-1
107-18-6
107-05-1
71-43-2
100-44-7
598-31-2
74-97-5
75-27-4
460-00-4
75-25-2
74-83-9
71-36-3
78-93-3
75-15-0
56-23-5
302-17-0
108-90-7
126-99-8
124-48-1
75-00-3
107-07-3
505-60-2
110-75-8
67-66-3
74-87-3
126-99-8
PP
PP
PP
PP
ht
a
a
a
PP
a
a
a
a
a
ht
PP
PP
a
PP
a
a
a
a
PP
PP
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
pc
                                   8260A -  1
                               Revision  1
                           September 1994

-------
   Appropriate Technique
Analyte
3-Chloropropene
3-Chloropropionitrile
1 , 2-Dibromo-3-chl oropropane
1,2-Dibromoethane
Dibromomethane
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
cis-l,4-Dichloro-2-butene
trans -l,4-Dichloro-2-butene
Di chl orodi f 1 uoromethane
1,1-Dichloroethane
1,2-Dichloroethane
1,1-Dichloroethene
trans -1,2-Di chl oroethene
1,2-Dichloropropane
l,3-Dichloro-2-propanol
cis-l,3-Dichloropropene
trans-l,3-Dichloropropene
1,2,3,4-Diepoxybutane
Di ethyl ether
1,4-Difluorobenzene (I.S.)
1,4-Dioxane
Epichlorohydrin
Ethanol
Ethyl acetate
Ethyl benzene
Ethylene oxide
Ethyl methacrylate
Hexachl orobutad i ene
Hexachloroethane
2-Hexanone
2-Hydroxypropionitrile
lodomethane
Isobutyl alcohol
Isopropyl benzene
Malononitrile
Methacrylonitrile
Methanol
Methylene chloride (DCM)
Methyl methacrylate
4-Methyl-2-pentanone (MIBK)
Naphthalene
Nitrobenzene
2-Nitropropane
CAS No.b
107-05-1
542-76-7
96-12-8
106-93-4
74-95-3
95-50-1
541-73-1
106-46-7
1476-11-5
110-57-6
75-71-8
75-34-3
107-06-2
75-35-4
156-60-5
78-87-5
96-23-1
10061-01-5
10061-02-6
1464-53-5
60-29-7
540-36-3
123-91-1
106-89-8
64-17-5
141-78-6
100-41-4
75-21-8
97-63-2
87-68-3
67-72-1
591-78-6
78-97-7
74-88-4
78-83-1
98-82-8
109-77-3
126-98-7
67-56-1
75-09-2
80-62-6
108-10-1
91-20-3
98-95-3
79-46-9
Purge-and-Trap
a
i
PP
a
a
a
a
a
a
PP
a
a
a
a
a
a
PP
a
a
a
a
a
PP
i
i
i
a
PP
a
a
i
PP
i
a
PP
a
PP
PP
i
a
a
PP
a
a
a
Direct
Injection
a
pc
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
pc
a
a
a
a
a
a
a
a
a
a
a
a
8260A - 2
    Revision 1
September 1994

-------
AocroDriate Techniaue

Analyte
Pentachloroethane
2-Picol ine
Propargyl alcohol
B-Propiolactone
Propionitrile (ethyl cyanide)
n-Propylamine
Pyridine
Styrene
1,1,1 , 2-Tetrachl oroethane
1 , 1 , 2, 2-Tetrachl oroethane
Tetrachloroethene
Toluene
1 , 2 , 4-Tri chl orobenzene
1,1,1-Trichloroethane
1,1,2-Trichloroethane
Trichloroethene
Trichlorofluoromethane
1,2,3-Trichloropropane
Vinyl acetate
Vinyl chloride
o-Xylene
m-Xylene
p-Xylene
a Adequate response by thi

CAS No.b
76-01-7
109-06-8
107-19-7
57-57-8
107-12-0
107-10-8
110-86-1
100-42-5
630-20-6
79-34-5
127-18-4
108-88-3
120-82-1
71-55-6
79-00-5
79-01-6
75-69-4
96-18-4
108-05-4
75-01-4
95-47-6
108-38-3
106-42-3
s technique.

Purge-and-Trap
i
PP
PP
PP
ht
a
i
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a

Direct
Injection
a
a
a
a
pc
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a

b Chemical Abstract Services Registry Number.
ht Method analyte only when
i Inappropriate technique
purged at 80°C
for this analyte.




pc Poor chromatographic behavior.
pp Poor purging efficiency
surr Surrogate
I.S. Internal Standard
resulting in high


EQLs.





      1.2   Method 8260 can be used to quantitate most volatile organic compounds
that have boiling points below 200°C and that are insoluble or slightly soluble
in water.  Volatile water-soluble compounds' can be included in this analytical
technique.  However,  for the more soluble compounds,  quantitation  limits are
approximately  ten  times  higher because  of poor  purging  efficiency.    Such
compounds  include  low-molecular-weight  halogenated hydrocarbons,  aromatics,
ketones,  nitriles,  acetates,  acrylates,  ethers,  and  sulfides.  See Tables 1 and
2 for lists of analytes and retention tines that have been  evaluated on a purge-
                                  8260A  - 3                         Revision 1
                                                                September 1994

-------
and-trap  GC/MS  system.    Also,  the method  detection  limits for  25  ml sample
volumes are presented.  The following  analytes are also  amenable to analysis by
Method 8260:

      Bromobenzene                  1-Chlorohexane
      n-Butylbenzene                2-Chlorotoluene
      sec-Butyl benzene              4-Chlorotoluene
      tert-Butylbenzene             Crotonaldehyde
      Chloroacetonitrile            Dibromofluoromethane
      1-Chlorobutane                cis-l,2-Dichloroethene
      1,3-Dichloropropane           Methyl-t-butyl  ether
      2,2-Dichloropropane           Pentafluorobenzene
      1,1-Dichloropropene           n-Propylbenzene
      Fluorobenzene                 1,2,3-Trichlorobenzene
      p-1sopropyltoluene            1,2,4-Trimethyl benzene
      Methyl acrylate               1,3,5-Trimethylbenzene

      1.3   The   estimated  quantitation  limit  (EQL)  of  Method  8260  for  an
individual compound is somewhat instrument dependent. Using  standard quadrupole
instrumentation,  limits  should  be approximately  5  M9/kg  (wet  weight)  for
soil/sediment samples, 0.5 mg/kg  (wet  weight)  for  wastes, and 5 fj.g/1 for ground
water (see Table 3).  Somewhat lower  limits may be achieved using an ion trap
mass spectrometer or other instrumentation of  improved design.  No matter which
instrument is used, EQLs will  be  proportionately higher for  sample extracts and
samples that require dilution  or  reduced sample  size to avoid saturation of the
detector.

      1.4   Method  8260  is based upon a purge-and-trap, gas chromatographic/mass
spectrometric (GC/MS) procedure.   This method  is restricted  to use by, or under
the supervision of,  analysts experienced in the use of purge-and-trap systems and
gas chromatograph/mass spectrometers,  and  skilled  in the interpretation of mass
spectra and their use as a quantitative tool.

      1.5   An  additional method for  sample introduction  is direct  injection.
This  technique  has  been  tested  for  the  analysis  of  waste oil  diluted  with
hexadecane  1:1  (vol/vol) and may  have application for  the analysis  of  some
alcohols and aldehydes  in aqueous samples.


2.0   SUMMARY OF METHOD

      2.1   The  volatile compounds  are introduced  into the gas chromatograph by
the purge-and-trap  method or by  direct  injection  (in  limited  applications).
Purged  sample  components are trapped in'a  tube   containing suitable sorbent
materials.  When purging is complete, the sorbent tube is heated and backflushed
with  helium  to  desorb  trapped sample components.  The analytes  are  desorbed
directly to  a  large  bore capillary or cryofocussed on a  capillary  precolumn
before being flash  evaporated to a narrow bore capillary  for analysis.   The
column is  temperature programmed to separate the analytes which are then detected
with a mass spectrometer  (MS) interfaced  to the gas chromatograph.   Wide  bore
capillary columns require a jet separator,  whereas  narrow bore capillary columns
can be directly interfaced to the ion  source.
                                   8260A  - 4                         Revision 1
                                                                September 1994

-------
      2.2   If the above sample introduction  techniques  are  not applicable,  a
portion of the sample is dispersed in solvent to dissolve the volatile organic
constituents.  A portion of the solution is combined with organic-free reagent
water  in  the  purge  chamber.   It  is  then  analyzed  by  purge-and-trap  GC/MS
following the normal  water method.

      2.3   Analytes  eluted from the capillary  column  are introduced into the
mass spectrometer via a  jet separator or a direct connection.  Identification of
target analytes is accomplished by comparing their mass spectra with the electron
impact (or electron impact-like) spectra of authentic standards.  Quantitation
is accomplished by comparing the response of a major  (quantitation) ion relative
to an internal standard with a five-point calibration curve.

      2.4   The method includes specific calibration and quality control  steps
that replace the general requirements in Method 8000.


3.0   INTERFERENCES

      3.1   Major contaminant  sources  are volatile  materials  in  the  laboratory
and impurities in the  inert purging gas and  in  the sorbent trap.   The use of non-
polytetrafluoroethylene  (PTFE)  thread  sealants,   plastic  tubing,   or  flow
controllers with rubber components should be  avoided since such materials out-gas
organic compounds  which will  be  concentrated  in  the  trap  during   the  purge
operation.  Analyses  of calibration and reagent blanks provide information about
the presence  of  contaminants.   When potential  interfering peaks  are noted in
blanks, the  analyst   should  change the  purge gas  source and  regenerate  the
molecular sieve  purge gas filter  (Figure  1).   Subtracting  blank values  from
sample results is not permitted.  If reporting values not corrected  for blanks
result in what the laboratory feels is  a  false positive for a sample, this should
be fully explained in text accompanying the uncorrected data.

      3.2   Interfering  contamination  may  occur  when  a  sample  containing  low
concentrations of volatile  organic  compounds is analyzed  immediately  after  a
sample  containing  high  concentrations  of  volatile organic  compounds.    The
preventive technique  is rinsing of the purging apparatus and sample syringes with
two portions of organic-free reagent water between  samples.   After analysis of
a sample containing  high concentrations  of volatile organic  compounds,  one or
more calibration blanks  should be analyzed to check for cross contamination.  For
samples containing large amounts of water soluble materials,  suspended solids,
high boiling compounds or high  concentrations  of compounds being determined, it
may be necessary to wash the  purging device  with a soap solution, rinse it with
organic-free reagent  water,  and then dry the purging device in an oven at 105°C.
In extreme situations, the whole purge and "trap device may require dismantling
and cleaning.  Screening of the samples prior to purge and trap  GC/MS analysis
is highly recommended  to prevent contamination of the system.   This is especially
true for soil  and waste samples. Screening may be accomplished with an automated
headspace technique or  by Method  3820  (Hexadecane  Extraction  and  Screening of
Purgeable Organics).

            3.2.1  The low  purging efficiency  of  many analytes from a 25  ml
      sample often results  in significant concentrations remaining in  the sample
      purge vessel after analysis.  After removal of  the analyzed sample aliquot


                                   8260A -  5                        Revision  1
                                                                September 1994

-------
      and  three  rinses of  the  purge  vessel  with analyte  free  water,  it  is
      required that the empty vessel  be subjected  to a heated purge cycle prior
      to the  analysis  of another  sample in the  same  purge vessel  to reduce
      sample to sample carryover.

      3.3   Special  precautions must be taken to analyze for methylene chloride.
The analytical and sample storage  area should be isolated from all  atmospheric
sources of methylene chloride.  Otherwise random background levels will result.
Since  methylene   chloride   will   permeate  through  PTFE  tubing,   all   gas
chromatography carrier gas lines and  purge  gas  plumbing  should  be  constructed
from stainless steel or copper tubing.   Laboratory clothing worn by the analyst
should be clean since  clothing previously exposed  to methylene  chloride fumes
during   liquid/liquid   extraction   procedures   can   contribute  to   sample
contamination.

      3.4   Samples  can  be   contaminated by  diffusion  of  volatile  organics
(particularly methylene chloride and fluorocarbons) through the septum seal into
the sample during  shipment and storage.  A trip blank prepared from organic-free
reagent water and carried through  the sampling and handling  protocol  can serve
as a check on such contamination.

      3.5   Use of sensitive mass spectrometers  to  achieve  lower detection level
will increase the  potential to detect  laboratory contaminants as  interferences.

      3.6   Direct  injection - Some  contamination may be  eliminated by baking out
the  column  between analyses.   Changing  the  injector liner  will  reduce  the
potential for cross-contamination.  A  portion of the analytical column may need
to be removed  in the case of extreme contamination.  Use  of  direct  injection will
result in the need for more  frequent  instrument maintenance.

      3.7   If hexadecane is added  to  samples or petroleum  samples are analyzed,
some  chromatographic  peaks  will  elute  after  the  target  analytes.    The  oven
temperature program must include a  post-analysis bake out  period to ensure that
semi-volatile hydrocarbons are volatilized.


4.0   APPARATUS AND MATERIALS

      4.1   Purge-and-trap device - aqueous  samples, described in Method  5030.

      4.2   Purge-and-trap device - solid samples,  described in Method  5030.

      4.3   Injection port liners (HP  catalogue  #18740-80200, or equivalent) are
modified for  direct  injection analysis  by  placing  a  1-cm plug  of pyrex  wool
approximately 50-60 mm  down the length
of  the  injection   port  towards  the
oven.  An 0.53 mm  ID column is mounted    septum       so — eo     o
1 cm into the liner from the oven side                      "^
of  the  injection  port,  according  to
manufacturer's  specifications.                       Modified Injector
                                  8260A  -  6                         Revision 1
                                                                September 1994

-------
4.4   Gas  chromatography/mass  spectrometer/data  system

      4.4.1  Gas chromatograph  - An  analytical   system  complete  with  a
temperature-programmable  gas   chromatograph  suitable   for  splitless
injection or interface to purge-and-trap apparatus.  The system includes
all  required  accessories,  including  syringes,   analytical  columns,  and
gases.  The  GC  should be equipped with  variable constant differential flow
controllers so that the column flow rate will remain constant throughout
desorption  and   temperature   program  operation.    For   some  column
configurations, the column oven must  be cooled  to < 30°C,  therefore,  a
subambient oven controller may be required.  The capillary column should
be directly coupled to the source.

            4.4.1.1     Capillary precolumn interface when using cryogenic
      cooling  - This device interfaces the purge  and trap concentrator to
      the  capillary  gas chromatograph.    The  interface condenses  the
      desorbed  sample components and focuses them into a narrow band on an
      uncoated  fused  silica  capillary precolumn.   When  the  interface is
      flash  heated, the  sample  is transferred to  the analytical  capillary
      column.

                  4.4.1.1.1  During   the    cryofocussing    step,   the
            temperature of the  fused silica in the interface  is maintained
            at -150°C  under  a  stream of  liquid  nitrogen.   After  the
            desorption period,  the  interface must be capable  of rapid
            heating to  250°C   in  15  seconds or  less  to   complete  the
            transfer  of analytes.

      4.4.2  Gas chromatographic columns

            4.4.2.1     Column 1  -  60 m  x 0.75 mm  ID  capillary column
      coated with  VOCOL  (Supelco), 1.5 p,m film thickness, or equivalent.

            4.4.2.2     Column 2 -  30  - 75 m x 0.53 mm ID capillary column
      coated with  DB-624 (J&W  Scientific),  Rtx-502.2  (RESTEK),  or VOCOL
      (Supelco), 3 /zm  film  thickness,  or  equivalent.

            4.4.2.3     Column 3  -  30 m x  0.25 - 0.32  mm  ID  capillary
      column coated with 95% dimethyl  - 5%  diphenyl  polysiloxane (DB-5,
      Rtx-5, SPB-5, or equivalent),  1 ^m film thickness.

            4.4.2.4     Column 4  -  60 m  x  0.32 mm  ID  capillary column
      coated with   DB-624  (J&W  Scientific),  1.8 fj.m  film thickness,  or
      equivalent.

      4.4.3  Mass  spectrometer   - Capable of  scanning from 35 to  300  amu
every 2  sec or less,  using  70 volts  (nominal)  electron energy  in  the
electron impact ionization mode. The mass spectrometer must be capable of
producing a mass  spectrum for  p-Bromofluorobenzene (BFB)  which  meets  all
of the criteria in Table 4 when 5-50  ng of the GC/MS tuning standard (BFB)
is  injected through  the  GC.    To  ensure  sufficient  precision  of  mass
spectral data,  the desirable MS scan rate allows acquisition of at least
five spectra while a  sample  component  elutes from the GC.


                            8260A - 7                         Revision  1
                                                          September 1994

-------
                  4.4.3.1      The  ion  trap mass  spectrometer may be used if it
            is  capable  of axial modulation to reduce ion-molecule reactions and
            can produce electron impact-like  spectra  that match those  in  the
            EPA/NIST Library.   In an ion trap  mass  spectrometer,  because ion-
            molecule reactions with water and  methanol may produce interferences
            that  coelute  with chloromethane and chloroethane, the base peak for
            both  of these analytes  will  be  at m/z 49.   This ion should be used
            as  the  quantitation ion in this  case.   The mass  spectrometer must be
            capable of  producing  a mass  spectrum  for BFB which meets all  of the
            criteria in Table 3 when  5  or 50  ng are  introduced.

            4.4.4 GC/MS interface - Two alternatives are used to interface the
      GC to the mass spectrometer.

                  4.4.4.1      Direct coupling by  inserting the column into the
            mass  spectrometer is  generally  used for  0.25-0.32 mm  id  columns.

                  4.4.4.2     A  separator including an all-glass transfer line
            and glass  enrichment  device or  split interface is  used with  an
            0.53  mm column.

                  4.4.4.3     Any enrichment device  or transfer line  can be used
            if  all  of the  performance specifications  described   in  Sec.  8
            (including  acceptable calibration at 50 ng or less) can be achieved.
            GC-to-MS interfaces constructed entirely of glass or of glass-lined
            materials are recommended.   Glass can be deactivated  by  silanizing
            with  dichlorodimethylsilane.

            4.4.5 Data system  -  A  computer system that allows  the  continuous
      acquisition  and  storage  on machine-readable  media  of all mass  spectra
      obtained throughout the duration  of  the  chromatographic  program must  be
      interfaced to the mass spectrometer.  The computer must have software that
      allows searching any  GC/MS data   file  for  ions  of a specified mass  and
      plotting such  ion  abundances  versus  time or scan number.  This type  of
      plot is defined as  an Extracted Ion Current  Profile (EICP).  Software must
      also  be  available  that  allows  integrating the  abundances  in any  EICP
      between specified time or scan-number limits.  The most recent version of
      the EPA/NIST Mass Spectral  Library should also be available.

      4.5   Microsyringes -  10, 25,  100, 250,  500, and  1,000 /xL.

      4.6   Syringe valve - Two-way, with Luer ends (three each),  if  applicable
to the purging device.

      4.7   Syringes -  5,  10, or  25 ml,  gas-tight  with shutoff valve.

      4.8   Balance - Analytical, 0.0001 g, and top-loading, 0.1 g.

      4.9   Glass scintillation vials -  20  mL,  with  Teflon  lined  screw-caps  or
glass culture tubes with  Teflon lined screw-caps.

      4.10  Vials - 2 mL,  for GC  autosampler.
                                  8260A  - 8                         Revision 1
                                                                September 1994

-------
      4.11  Disposable pipets -  Pasteur.

      4.12  Volumetric flasks, Class A  -  10 ml  and  100  ml,  with ground-glass
stoppers.

      4.13  Spatula -  Stainless  steel.


5.0   REAGENTS

      5.1   Reagent grade inorganic chemicals shall  be used in all  tests.  Unless
otherwise indicated, it is  intended that all inorganic reagents  shall conform to
the  specifications of the  Committee on  Analytical  Reagents  of  the  American
Chemical Society, where such specifications are  available.  Other grades may be
used, provided it is first  ascertained  that the  reagent is of sufficiently high
purity to permit its use without lessening the accuracy of the determination.

      5.2   Organic-free  reagent water  - All references to water in this method
refer to organic-free reagent water,  as defined  in Chapter One.

      5.3   Methanol,  CH3OH  - Pesticide quality  or equivalent,  demonstrated to
be free of analytes.  Store  apart from other solvents.

      5.4   Reagent Hexadecane - Reagent hexadecane is defined  as hexadecane in
which interference is  not observed at the method  detection limit of compounds of
interest.

            5.4.1  In order to demonstrate  that  all  interfering volatiles have
      been  removed from  the hexadecane,  a  direct  injection  blank  must  be
      analyzed.

      5,5   Polyethylene  glycol,  H(OCH2CH2),,OH  -  Free of  interferences  at  the
detection limit of the target analytes.

      5.6   Hydrochloric  acid (1:1  v/v),  HC1  - Carefully  add a measured volume
of concentrated HC1 to an equal  volume of organic-free reagent water.

      5.7   Stock solutions - Stock solutions may be prepared from pure standard
materials or purchased as certified solutions.  Prepare stock standard solutions
in methanol, using assayed liquids or gases, as  appropriate.

            5.7.1  Place about 9.8  ml  of methanol  in a 10 ml tared ground-glass-
      stoppered volumetric flask.  Allow  the  flask  to  stand,  unstoppered,  for
      about 10 minutes or until  all  alcohol-wetted surfaces have dried.   Weigh
      the flask to the nearest 0.0001 g.

            5.7.2  Add  the assayed  reference material, as  described  below.

                  5.7.2.1      Liquids - Using a  100 /uL syringe, immediately add
            two or  more drops of  assayed  reference  material  to  the  flask;  then
            reweigh.   The liquid must  fall  directly into the  alcohol  without
            contacting the  neck of the  flask.
                                  8260A  -  9                         Revision 1
                                                                September 1994

-------
            5.7.2.2      Gases  - To prepare  standards  for any compounds
      that boil below 30°C (e.g. bromomethane,  chloroethane, chloromethane,
      or  vinyl  chloride), fill  a 5 mL  valved gas-tight syringe with the
      reference standard to the  5.0  ml mark.  Lower the  needle  to 5 mm
      above the methanol meniscus. Slowly introduce the  reference standard
      above the surface of the liquid.  The heavy gas will rapidly dissolve
      in  the  methanol.   Standards may also be prepared  by using a lecture
      bottle  equipped  with  a Hamilton  Lecture Bottle  Septum (#86600).
      Attach  Teflon tubing to the side arm relief valve and  direct a gentle
      stream  of gas  into the methanol meniscus.

      5.7.3  Reweigh,  dilute to volume,  stopper, and then mix by inverting
the flask several times.   Calculate  the concentration  in  milligrams per
liter (mg/L)  from the net gain in weight.  When  compound purity is assayed
to  be 96% or  greater,  the weight  may be  used  without  correction  to
calculate the concentration of the  stock standard.  Commercially prepared
stock standards may be  used at any  concentration if they are certified by
the manufacturer or by an independent source.

      5.7.4  Transfer the  stock  standard solution  into a  bottle  with  a
Teflon lined screw-cap.  Store,  with  minimal  headspace, at -10°C to  -20°C
and protect from light.

      5.7.5  Prepare  fresh  standards   for  gases   weekly  or   sooner  if
comparison with check standards  indicates a problem.  Reactive compounds
such as 2-chloroethyl vinyl  ether and styrene may need to be prepared more
frequently.   All  other standards must   be replaced  after  six  months,  or
sooner if comparison with check  standards indicates a problem.  Both gas
and  liquid  standards  must  be  monitored  closely  by comparison  to the
initial calibration curve and  by comparison to QC check  standards.   It may
be  necessary to  replace the  standards  more  frequently if  either check
exceeds a 20% drift.

      5.7.6  Optionally  calibration  using a certified gaseous mixture can
be  accomplished  daily  utilizing commercially  available gaseous  analyte
mixture  of  bromomethane,  chloromethane,  chloroethane, vinyl  chloride,
dichlorodifluoromethane  and trichlorofluoromethane  in nitrogen.  These
mixtures of  documented  quality are  stable  for  as  long  as  six  months
without refrigeration.   (VOA-CYL  III, RESTEK  Corporation,  Cat. #20194 or
equivalent).

            5.7.6.1      Preparation  of Calibration Standards  From a Gas
      Mixture

                  5.7.6.1.1  Before  removing the  cylinder shipping cap,
            be  sure the valve  is completely closed  (turn clockwise).  The
            contents are  under  pressure and should  be used  in  a well-
            ventilated  area.

                  5.7.6.1.2  Wrap the pipe thread  end of the Luer fitting
            with  Teflon tape.  Remove the shipping cap from the cylinder
            and replace it with the Luer fitting.
                            8260A - 10                        Revision 1
                                                          September 1994

-------
       5.7.6.1.3   Transfer half the working standard containing
other  analytes,  internal standards,  and  surrogates  to the
purge apparatus.

       5.7.6.1.4   Purge  the  Luer fitting and stem on the gas
cylinder prior to  sample  removal using the following sequence:

       a)     Connect  either  the  100 /uL or 500 jxL  Luer syringe
             to the inlet fitting of  the cylinder.

       b)     Make  sure the on/off valve on the syringe  is in
             the open position.

       c)     Slowly  open the   valve  on   the  cylinder  and
             withdraw a  full  syringe  volume.

       d)     Be sure  to close the valve on the cylinder before
             you withdraw the syringe from the  Luer  fitting.

       e)     Expel  the  gas   from  the  syringe  into  a  well-
             ventilated  area.

       f)     Repeat steps a  through  e one more time to fully
             purge the fitting.

       5.7.6.1.5   Once the fitting and stem have  been purged,
quickly withdraw  the volume of gas you  require  using  steps
5.6.6.1.4(a) through  (d).  Be sure to close the  valve on the
cylinder and syringe before  you  withdraw the syringe from the
Luer fitting.

       5.7.6.1.6   Open the syringe on/off valve for 5 seconds
to reduce the syringe pressure to atmospheric pressure.   The
pressure in the cylinder is  -30 psi.

       5.7.6.1.7   The gas mixture should be quickly transferred
into the reagent water through the female Luer fitting located
above the purging  vessel.

       NOTE:  Make   sure   the  arrow  on  the 4-way  valve  is
             pointing toward  the female  Luer fitting  when
             transferring the sample from the syringe. Be sure
             to switch the  4-way valve  back  to  the  closed
             position before removing  the syringe  from the
             Luer  fitting.

       5.7.6.1.8   Transfer the  remaining half  of  the working
standard into the  purging vessel.  This procedure  insures that
the  total  volume   of gas mix  is flushed  into  the  purging
vessel, with none  remaining  in the valve or lines.

       5.7.6.1.9   Concentration   of   each  compound   in   the
cylinder is typically 0.0025
                8260A - 11                        Revision 1
                                              September 1994

-------
                        5.7.6.1.10 The fol 1 owi ng are the recommended gas vol umes
                  spiked  into  5  ml  of  water  to  produce  a  typical  5-point
                  calibration:

                        Gas                     Calibration
                        Volume                  Concentration
                         40 ni                      20 M9/L
                        100 Hi                      50 M9/L
                        200 fti                     100 M9/L
                        300 ML                     150 M9/L
                        400 ML                     200 M9/L

                        5.7.6.1.11 The following are the recommended gas volumes
                  spiked  into  25 ml  of  water  to produce  a  typical  5-point
                  calibration:

                        Gas                     Calibration
                        Volume                  Concentration
                         10 /LtL                       1 M9/L
                         20 ML                       2 M9/L
                         50 ML                       5 M9/L
                        100 ML                      10 M9/L
                        250 ML                      25 M9/L

      5.8   Secondary  dilution  standards -  Using  stock standard  solutions,
prepare in methanol, secondary dilution  standards  containing the compounds  of
interest,  either singly  or mixed together. Secondary dilution standards must be
stored with  minimal  headspace  and should be  checked  frequently  for  signs  of
degradation  or  evaporation, especially  just prior  to  preparing calibration
standards from them.  Store in a vial  with no headspace for one week only.

      5.9   Surrogate  standards  - The  surrogates  recommended  are toluene-d8,
4-bromofluorobenzene,  l,2-dichloroethane-d4,  and dibromofluoromethane.   Other
compounds may be used as surrogates,  depending upon the analysis requirements.
A stock surrogate solution in methanol should be prepared as described above, and
a surrogate  standard spiking  solution  should  be prepared from  the  stock at a
concentration of 50-250 M9/10 niL in methanol.   Each water  sample undergoing
GC/MS analysis must be spiked with 10 ML of the surrogate  spiking solution prior
to analysis.

            5.9.1  If a  more sensitive mass spectrometer is  employed to achieve
      lower detection levels,  more dilute surrogate solutions may be required.

      5.10  Internal  standards   -   The  recommended   internal   standards  are
fluorobenzene, chlorobenzene-d5, and  l,4-dichlorobenzene-d4.  Other compounds may
be used as  internal standards as long as they have retention times similar  to the
compounds being  detected by GC/MS. Prepare internal  standard  stock  and secondary
dilution  standards  in methanol  using the procedures described in Sees.  5.7 and
5.8.  It  is recommended  that the  secondary dilution standard should be prepared
at a concentration  of 25 mg/L of each internal  standard compound.  Addition  of
10 ML of this standard  to 5.0  ml  of sample or calibration standard would  be the
equivalent of 50 jtzg/L.
                                  8260A - 12                        Revision 1
                                                                September 1994

-------
            5.10.1       If  a  more sensitive mass  spectrometer  is employed to
      achieve lower detection  levels,  more dilute internal standard solutions
      may be  required.   Area  counts of the  internal  standard peaks should be
      between 50-200% of the  area  of the  target  analytes  in  the mid-point
      calibration analysis.

      5.11  4-Bromofluorobenzene (BFB)  standard - A standard solution containing
25 ng//iL of BFB  in methanol should be  prepared.

            5.11.1       If  a  more sensitive mass  spectrometer  is employed to
      achieve lower detection  levels, a more dilute BFB standard  solution may be
      required.

      5.12  Calibration  standards -  Calibration standards at a minimum of five
concentrations should  be prepared from  the secondary dilution of  stock standards
(see Sees.  5.7 and 5.8).  Prepare these solutions in organic-free reagent water.
One of  the  concentrations should be  at  a concentration near,  but above, the
method detection limit.   The remaining concentrations should correspond to the
expected range of concentrations found  in real samples but should not exceed the
working range  of  the GC/MS system. Each standard should contain each analyte for
detection by  this method.   It is EPA's intent that  all  target  analytes  for a
particular analysis be included in the  calibration  standard(s).  However, these
target analytes  may  not  include  the entire List of Analytes  (Sec. 1.1) for which
the method has been demonstrated.   However,  the  laboratory shall  not report a
quantitative result  for  a target analyte that  was not included in  the calibration
standard(s).  Calibration standards must be prepared daily.

      5.13  Matrix spiking  standards  -  Matrix  spiking standards should  be
prepared from volatile  organic compounds  which will  be  representative  of the
compounds being  investigated.  At a minimum, the matrix spike should include 1,1-
dichloroethene,   trichloroethene,  chlorobenzene,  toluene, and  benzene.   It is
desirable to  perform  a  matrix  spike  using  compounds found in  samples.   Some
permits may require spiking specific compounds of interest, especially if they
are polar and would  not  be  represented  by  the  above listed  compounds.   The
standard should  be prepared  in  methanol,   with  each compound  present  at  a
concentration of 250 jug/10.0 ml.

            5.13.1       If  a  more sensitive mass  spectrometer  is employed to
      achieve lower detection levels, more dilute matrix spiking solutions may
      be required.

      5.14  Great care must  be taken  to maintain the integrity of all  standard
solutions.   It is recommended  all standards  in methanol  be stored at -10°C to
-20°C  in amber bottles with  Teflon lined screw-caps.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See the introductory material to this chapter,  Organic Analytes, Sec.
4.1.
                                  8260A - 13                        Revision 1
                                                                September 1994

-------
7.0   PROCEDURE

      7.1   Three  alternate  methods  are provided for sample introduction.  All
internal standards,  surrogates, and matrix spikes (when applicable) must be added
to samples before introduction.

            7.1.1  Direct  injection   -   in  very limited  application,  (e.g.,
      volatiles in  waste  oil  or  aqueous  process wastes) direct  injection of
      aqueous  samples  or  samples diluted  according  to  Method  3585 may be
      appropriate.  Direct injection has been used for the analysis of volatiles
      in waste  oil  (diluted 1:1  with  hexadecane)  and for  determining if the
      sample is ignitable  (aqueous  injection,  Methods 1010 or  1020).   Direct
      injection is  only permitted for  the  determination of  volatiles at the
      toxicity  characteristic  (TC)  regulatory  limits,  at concentrations in
      excess of 10,000 M9/U or for water-soluble compounds that do not purge.

            7.1.2  Purge-and-trap  for  aqueous  samples,  see  Method  5030  for
      details.

            7.1.3  Purge-and-trap for solid samples, see Method  5030 for details.

      7.2   Recommended  Chromatographic  conditions

            7.2.1  General:

                  Injector  temperature:          200-225°C
                  Transfer  line temperature:    250-300°C

            7.2.2  Column 1  (A  sample  chromatogram is presented in Figure  5)

                  Carrier gas  (He) flow rate:    15 mL/min
                  Initial temperature:           10°C,  hold  for 5 minutes
                  Temperature  program:           6°C/min to  160°C
                  Final  temperature:             160°C,  hold  until  all  expected
                                                compounds have eluted.

            7.2.3  Column 2,  Cryogenic  cooling  (A  sample  chromatogram  is
      presented in Figure  6)

                  Carrier gas  (He) flow rate:    15 mL/min
                  Initial temperature:           10°C,  hold  for 5 minutes
                  Temperature  program:           6°C/min to  160°C
                  Final  temperature:             160°C,  hold  until  all  expected
                                                compounds have eluted.

            7.2.4  Column 2, Non-cryogenic cooling  (A sample chromatogram is
      presented in Figure 7).  It is recommended that carrier gas flow and split
      and make-up  gases  be set  using performance  of  standards as guidance.  Set
      the carrier gas head  pressure to <=  10 psi  and the  split to  « 30 mL/min.
      Optimize the make-up gas  flow for the  separator (approximately 30 mL/min)
      by injecting BFB,  and determining the optimum response  when varying the
      make-up gas.  This will  require  several  injections of  BFB.   Next,  make
      several injections of the volatile working standard with all analytes of

                                  8260A -  14                        Revision 1
                                                                September 1994

-------
interest.  Adjust the carrier and split  to  provide optimum chromatography
and response.  This is an  especially  critical  adjustment  for the volatile
gas analytes.  The head pressure should  optimize  between  8-12 psi and the
split between  20-60 mL/min.   The use  of  the splitter  is  important to
minimize the effect  of  water  on analyte response,  to allow the use of a
larger volume of helium during trap desorption, and to slow column flow.

      Initial temperature:     45°C, hold for 2 minutes
      Temperature program:     8°C/min to 200°C
      Final  temperature:       200°C, hold for 6 minutes.

      A  trap preheated to 150°C prior to trap  desorption is  required to
provide adequate chromatography of the gas analytes.

      7.2.5  Column 3 (A sample chromatogram is presented in Figure 8)

            Carrier gas (He) flow rate:    4 mL/min
            Initial temperature:          10°C,  hold for 5 minutes
            Temperature program:          6°C/min to 70°C, then  15°C/min
                                          to 145°C
            Final temperature:            145°C,  hold until  all expected
                                          compounds have eluted.

      7.2.6  Direct injection - Column 2

            Carrier gas (He) flow rate:    4 mL/min
            Column:                       J&W DB-624, 70m x 0.53 mm
            Initial temperature:          40°C,  hold for 3 minutes
            Temperature program:          8°C/min
            Final temperature:            260°C,  hold until  all expected
                                          compounds have eluted.
            Column Bake out  (direct inj):  75 minutes
            Injector temperature:         200-225°C
            Transfer line  temperature:     250-300°C

      7.2.7  Direct Split Interface -  Column 4

            Carrier gas (He) flow rate:    1.5 mL/min
            Initial temperature:          35°C, hold for 2 minutes
            Temperature program:          4°C/min to 50°C
                                          10'C/min to 220'C
            Final temperature:            220°C,  hold until  all expected
                                          compounds have eluted
            Split ratio:                  100:1
            Injector temperature:         125°C

7.3   Initial calibration  -  the recommended  MS operating conditions

      Mass range:              35-260 amu
      Scan time:               0.6-2 sec/scan
      Source  temperature:     According  to manufacturer's specifications
                            8260A - 15                        Revision 1
                                                          September 1994

-------
      Ion trap only:           Set  axial modulation, manifold temperature,
                              and   emission  current  to  manufacturer's
                              recommendations

      7.3.1  Each GC/MS system must be hardware-tuned  to meet the criteria
in Table 4  for a  5-50  ng injection  or  purging of 4-bromofluorobenzene
(2 IJ.L injection of the BFB standard).  Analyses must not  begin until these
criteria are met.

      7.3.2  Set up the purge-and-trap system as outlined  in Method 5030 if
purge-and-trap  analysis   is  to  be  utilized.    A set of at least  five
calibration  standards containing  the method  analytes  is  needed.   One
calibration  standard should  contain  each analyte  at   a  concentration
approaching but greater than  the method detection limit (Table 1) for that
compound; the other  calibration  standards  should  contain  analytes  at
concentrations that define the range  of the method.  Calibration should be
done  using   the  sample  introduction technique  that will  be  used  for
samples.  For  Method 5030, the  purging efficiency for  5 ml  of water is
greater  than for  25 mL.   Therefore, develop  the  standard curve  with
whichever volume of  sample that  will be analyzed.

            7.3.2.1      To prepare a calibration standard for purge-and-
      trap  or aqueous direct  injection,  add an  appropriate  volume  of a
      secondary dilution  standard  solution to an aliquot of organic-free
      reagent water in a volumetric flask.   Use  a microsyringe and rapidly
      inject  the alcoholic standard into  the expanded area  of the filled
      volumetric flask.   Remove  the  needle  as  quickly as possible after
      injection.  Mix by inverting  the flask three times only.  Discard the
      contents contained  in the neck  of the flask.  Aqueous standards are
      not stable and  should be prepared daily.   Transfer  5.0 ml (or 25 ml
      if lower detection  limits  are  required)  of each standard to  a gas
      tight syringe along with 10  /xL  of internal  standard.  Then transfer
      the contents  to a purging device or syringe.  Perform purge-and-trap
      or direct injection as  outlined in Method 5030.

            7.3.2.2      To prepare  a calibration  standard  for  direct
      injection analysis  of oil, dilute standards in  hexadecane.

      7.3.3  Tabulate the  area response  of the  characteristic  ions  (see
Table  5) against  concentration  for each  compound  and each  internal
standard.  Calculate response factors (RF) for each compound relative to
one of  the  internal  standards.    The  internal  standard  selected  for the
calculation  of the RF for a compound  should be the internal  standard that
has a retention time  closest  to the compound being measured (Sec.  7.6.2).
The RF is calculated as follows:
                            8260A - 16                        Revision 1
                                                          September 1994

-------
                         RF  =  (AxCis)/(AisCx)

where:

      Ax     =     Area of the  characteristic ion for the compound being
                  measured.
      Ais     =     Area  of  the  characteristic  ion  for  the  specific
                  internal  standard.
      Cjs     =     Concentration of the specific  internal standard.
      Cx     =     Concentration of the compound  being measured.

      7.3.4  The  average  RF must  be  calculated and  recorded  for each
compound using the five  RF  values  calculated for each compound from the
initial (5-point) calibration curve.  A system performance check should be
made before this  calibration curve  is  used.   Five  compounds (the  System
Performance Check Compounds, or SPCCs) are checked for a minimum average
relative  response factor.    These  compounds  are  chloromethane;  1,1-
dichloroethane; bromoform;  1,1,2,2-tetrachloroethane; and chlorobenzene.
These compounds  are used to check  compound  instability  and to check for
degradation caused by contaminated  lines  or  active  sites in the system.
Examples of these occurrences are:

            7.3.4.1       Chloromethane  - This  compound is the most  likely
      compound to be  lost if the  purge  flow  is too  fast.

            7.3.4.2       Bromoform - This compound is  one  of the compounds
      most  likely to  be purged  very poorly if  the purge flow is too slow.
      Cold  spots  and/or  active  sites in the transfer lines may adversely
      affect  response.   Response  of the  quantitation ion  (m/z  173)  is
      directly affected  by  the  tuning  of   BFB  at   ions  m/z  174/176.
      Increasing  the m/z  174/176  ratio relative to  m/z 95  may improve
      bromoform response.

            7.3.4.3       Tetrachloroethane and 1,1-dichloroethane - These
      compounds are degraded by contaminated transfer lines in purge-and-
      trap  systems and/or active  sites  in  trapping materials.

      7.3.5  Using the  RFs   from the  initial   calibration,  calculate  and
record the percent relative standard deviation (%RSD) for all compounds.
The percent RSD is calculated as  follows:


                   % RSD =  -JH. x 100%
                            RFX
where:
      RSD   =     Relative standard deviation.
      RFX    =     mean of 5 initial RFs for a compound.
      SD    =     standard deviation of the 5 initial RFs for a compound.
                            8260A - 17                        Revision 1
                                                          September 1994

-------
                    SD =
                          A  (RFi-RF):
                                n-l
            where:

                  RFi    =  RF  for each of the 5 calibration levels
                  N      =  number of RF values (i.e.,  5)

      The percent relative  standard deviation should be less than  15% for
each  compound.   However, the  %RSD for each  individual Calibration Check
Compound  (CCC) must  be less than 30%.  The CCCs are:

      1,1-Dichloroethene,
      Chloroform,
      1,2-Dichloropropane,
      Toluene,
      Ethyl benzene,  and
      Vinyl  chloride.

            7.3.5.1       If a %RSD greater than 30  percent  is measured for
      any CCC,  then   corrective action  to  eliminate  a  system leak and/or
      column reactive sites is  required before reattempting calibration.

      7.3.6  Linearity -  If the %RSD of any compound is 15% or less, then
the  relative  response   factor  is  assumed   to   be   constant  over  the
calibration range, and the average  relative  response  factor may be used
for quantitation.

            7.3.6.1       If the %RSD of any compound is greater than 15%,
      construct   calibration   curves  of   area   ratio   (A/Ais)   versus
      concentration using first or higher order regression  fit of the five
      calibration points.  The  analyst should select the regression order
      which  introduces the  least  calibration  error into the quantitation.
      The use of calibration curves  is a recommended alternative to average
      response factor calibration (Sec. 7.6.2.4),  and  a useful diagnostic
      of standard preparation  accuracy and  absorption  activity  in  the
      chromatographic system.

      7.3.7  These curves  are verified each  shift by purging a performance
standard.   Recalibration is required  only  if calibration  and  on-going
performance criteria cannot be met.

7.4   GC/MS  calibration verification

      7.4.1  Prior to the  analysis of samples, inject or purge 5-50 ng of
the 4-bromofluorobenzene standard  following  Method  5030.   The  resultant
mass spectra for the BFB must meet  all  of  the criteria  given in Table 4
before sample analysis begins.  These criteria must be demonstrated each
12-hour shift.
                            8260A - 18                        Revision 1
                                                          September 1994

-------
      7.4.2 The initial calibration curve (Sec.  7.3)  for each compound of
interest must be checked and verified once every 12 hours during analysis
with the introduction technique used  for samples.  This is accomplished by
analyzing  a  calibration  standard  that is  at  a  concentration  near the
midpoint concentration for the working range of  the GC/MS by checking the
SPCC and CCC.

      7.4.3 System  Performance  Check  Compounds  (SPCCs)  -  A  system
performance check must  be made each  12 hours.   If the SPCC criteria are
met, a comparison of relative response factors is  made for all compounds.
This is the same check that  is  applied during the initial calibration.  If
the minimum  relative response  factors  are  not  met,   the  system  must be
evaluated, and  corrective  action  must be taken  before  sample analysis
begins.    Some  possible problems  are   standard mixture  degradation,
injection port inlet contamination, contamination  at  the front end of the
analytical column,  and  active sites  in  the column  or  chromatographic
system.

            7.4.3.1      The minimum relative response factor for volatile
      SPCCs are as  follows:

            Chloromethane                              0.10
            1,1-Dichloroethane                         0.10
            Bromoform                                 >0.10
            Chlorobenzene                              0.30
            1,1,2,2-Tetrachloroethane                  0.30

      7.4.4 Calibration  Check  Compounds  (CCCs)   -   After  the  system
performance check is met, CCCs listed in Sec. 7.3.5 are used to check the
validity of the initial calibration.

      Calculate the  percent  drift  using the  following equation:

                    % Drift  = (C, - CC)/C,  x  100

where:

      C, =   Calibration Check Compound standard  concentration.
      Cc =   Measured concentration using selected  quantitation  method.

      If the  percent  drift  for each  CCC  is  less  than 20%,  the  initial
calibration is assumed to be  valid.   If the  criterion is  not met (> 20%
drift),  for  any one  CCC,  corrective  action must be taken.    Problems
similar to those listed  under  SPCCs  could affect  this criterion.   If no
source of the problem can be determined after corrective  action has been
taken, a new  five  point calibration MUST be generated.   This  criterion
MUST be met before quantitative sample analysis  begins.  If the CCCs are
not required  analytes by the permit,  then  all required analytes  must meet
the 20% drift criterion.

      7.4.5 The internal  standard  responses  and  retention times  in the
check calibration standard must be evaluated immediately  after  or during
data acquisition.   If  the retention time for  any  internal standard changes


                            8260A - 19                        Revision  1
                                                          September 1994

-------
by more than 30 seconds  from  the  last  calibration check (12 hours), the
chromatographic system must be inspected  for malfunctions and corrections
must be made,  as  required.   If the EICP  area for any of  the internal
standards changes  by a factor of two (-50% to +100%) from the last daily
calibration check standard, the mass  spectrometer must be inspected for
malfunctions  and   corrections  must  be  made,  as  appropriate.    When
corrections are made,  reanalysis of samples analyzed while the system was
malfunctioning is  necessary.

7.5  GC/MS  analysis

     7.5.1  It is  highly  recommended  that  the extract  be  screened  on a
headspace-GC/FID  (Methods  3810/8015),  headspace-GC/PID/ELCD  (Methods
3810/8021), or waste dilution-GC/PID/ELCD  (Methods  3585/8021)  using the
same type of capillary column.  This  will  minimize  contamination  of the
GC/MS system from  unexpectedly high concentrations of organic compounds.
Use of  screening  is  particularly  important when  this  method  is used to
achieve low detection levels.

     7.5.2  All samples and standard solutions  must be  allowed to warm to
ambient temperature before analysis.  Set up the purge-and-trap  system as
outlined in Method 5030 if purge-and-trap introduction will  be used.

     7.5.3  BFB  tuning  criteria   and  GC/MS  calibration  verification
criteria must be met before analyzing samples.

            7.5.3.1      Remove the  plunger from a  5 ml  syringe and attach
     a  closed  syringe valve.  If lower detection limits  are required, use
     a  25 ml syringe.  Open the sample or standard bottle, which has been
     allowed to come to ambient temperature, and carefully  pour the sample
     into  the  syringe  barrel  to just  short of overflowing.   Replace the
     syringe plunger and compress the sample.   Open the  syringe  valve and
     vent  any  residual  air while  adjusting the sample  volume  to 5.0 ml.

     7.5.4  The process  of taking  an  aliquot  destroys the  validity of
aqueous and soil samples  for future  analysis; therefore, if there is only
one VGA vial,  the  analyst should prepare  a second  aliquot for analysis at
this time  to  protect against possible  loss  of sample  integrity.   This
second  sample  is  maintained only until  such  time when the  analyst has
determined that the first sample has been analyzed properly.  For aqueous
samples, filling  one  20 ml  syringe would  require  the use of  only one
syringe.   If  a second analysis  is  needed  from  a  syringe,  it must be
analyzed within 24 hours.  Care  must be taken to prevent air from leaking
into the syringe.

            7.5.4.1      The   following   procedure  is   appropriate   for
     diluting  aqueous  purgeable  samples.   All steps  must  be  performed
     without  delays  until  the diluted  sample  is in  a gas-tight  syringe.

                  7.5.4.1.1   Dilutions may be  made  in  volumetric  flasks
            (10 to 100 ml).  Select the volumetric flask that will  allow
            for the necessary  dilution.   Intermediate  dilutions may be
            necessary for extremely large dilutions.


                            8260A - 20                        Revision 1
                                                          September 1994

-------
                  7.5.4.1.2  Calculate the approximate volume of organic-
            free  reagent  water  to  be   added  to  the  volumetric  flask
            selected and add slightly less  than this quantity of organic-
            free reagent water to the flask.

                  7.5.4.1.3  Inject the proper aliquot of sample from the
            syringe into the flask.   Aliquots of less than 1 ml are not
            recommended.  Dilute  the  sample to the mark with organic-free
            reagent water.   Cap the flask,  invert, and shake three times.
            Repeat above procedure for additional dilutions.

                  7.5.4.1.4  Fill a 5 ml syringe  with  the diluted sample.

            7.5.4.2     Compositing   aqueous   samples  prior  to  GC/MS
      analysis

                  7.5.4.2.1  Add  5  ml or  equal  larger  amounts  of each
            sample (up to 5  samples are allowed)  to a 25 mL glass syringe.
            Special  precautions must be made to maintain  zero headspace in
            the syringe.

                  7.5.4.2.2  The samples must  be  cooled at 4°C during this
            step to minimize volatilization losses.

                  7.5.4.2.3  Mix  well  and  draw  out  a 5 ml  aliquot  for
            analysis.

                  7.5.4.2.4  Follow  sample  introduction,  purging,  and
            desorption steps described in Method  5030.

                  7.5.4.2.5  If  less than five  samples  are  used  for
            compositing, a  proportionately  smaller  syringe may  be used
            unless a 25 ml  sample is  to  be  purged.

      7.5.5  Add  10.0  /xL  of surrogate  spiking   solution  and  10  p.1  of
internal standard  spiking  solution  to each  sample.   The  surrogate  and
internal standards may  be mixed  and  added  as  a  single spiking  solution.
The addition of 10 /j,l of the surrogate spiking solution to 5 ml of sample
is equivalent to a concentration  of  50 /ig/L of  each  surrogate  standard.
The addition of 10 jul of the surrogate spiking solution to 5 g  of sample
is equivalent to a concentration  of 50 M9/kg of each  surrogate  standard.

            7.5.5.1      If  a more sensitive mass  spectrometer is employed
      to achieve lower detection levels, more dilute surrogate and internal
      standard  solutions may be required.

      7.5.6  Perform purge-and-trap or direct injection  by Method 5030.  If
the  initial  analysis  of  sample or  a   dilution of  the  sample  has  a
concentration of analytes that exceeds the  initial calibration range,  the
sample  must  be  reanalyzed  at   a  higher  dilution.    Secondary  ion
quantitation is allowed only when there are  sample interferences with the
primary ion.   When  a  sample is  analyzed that has saturated  ions  from a
compound,  this analysis must be  followed by a blank organic-free reagent


                            8260A -  21                        Revision 1
                                                          September 1994

-------
water analysis.  If the blank analysis is not free of interferences, the
system must be decontaminated.  Sample analysis may not resume until the
blank analysis is demonstrated to be free of interferences.

            7.5.6.1.     All  dilutions should  keep  the response  of the
      major constituents (previously saturated peaks) in the upper half of
      the  linear  range of the curve.  Proceed to Sees. 7.6.1 and 7.6.2 for
      qualitative and  quantitative  analysis.

      7.5.7 For  matrix  spike analysis,  add  10 juL  of  the matrix  spike
solution (Sec. 5.13) to  the 5 ml of sample to be purged.  Disregarding any
dilutions, this is  equivalent to a concentration of 50 jzg/L  of each matrix
spike standard.

7.6   Data interpretation

      7.6.1 Qualitative analysis

            7.6.1.1     The   qualitative  identification   of  compounds
      determined  by  this  method  is  based  on  retention  time,  and  on
      comparison  of the sample mass  spectrum,  after background correction,
      with characteristic ions in a reference mass spectrum.  The reference
      mass  spectrum  must  be  generated   by the  laboratory  using  the
      conditions  of  this  method.    The  characteristic  ions  from  the
      reference mass spectrum are defined  to be  the three ions of greatest
      relative intensity, or  any  ions over 30% relative intensity if less
      than three  such ions occur  in the  reference  spectrum.   Compounds
      should be identified as  present  when  the  criteria  below  are met.

                  7.6.1.1.1  The intensities of the characteristic ions
            of a  compound maximize  in  the same  scan or within one scan of
            each  other.   Selection of a peak  by  a data  system  target
            compound  search  routine where  the search  is  based on  the
            presence  of  a  target  chromatographic  peak containing  ions
            specific   for  the  target  compound  at  a  compound-specific
            retention  time  will  be  accepted as  meeting  this criterion.

                  7.6.1.1.2  The RRT  of  the sample component  is  within
            ± 0.06 RRT units  of the RRT of the  standard component.

                  7.6.1.1.3  The    relative    intensities    of    the
            characteristic   ions  agree  within  30%   of   the  relative
            intensities  of  these  ions   in   the   reference   spectrum.
            (Example:    For  an  ion' with  an  abundance  of  50%  in  the
            reference  spectrum, the corresponding abundance  in  a  sample
            spectrum  can range between 20% and  80%.)

                  7.6.1.1.4  Structural isomers that produce very similar
            mass spectra should  be identified   as individual  isomers  if
            they  have  sufficiently   different  GC   retention   times.
            Sufficient GC resolution  is  achieved  if  the  height  of the
            valley between two  isomer peaks is less than 25% of the sum of
                            8260A - 22                        Revision 1
                                                          September 1994

-------
       the  two  peak  heights.   Otherwise,  structural  isomers  are
       identified  as  isomeric  pairs.

             7.6.1.1.5  Identification  is  hampered  when   sample
       components  are not resolved chromatographically and  produce
       mass  spectra  containing  ions  contributed by  more than  one
       analyte.  When gas  chromatographic  peaks  obviously represent
       more  than one  sample component (i.e., a  broadened  peak  with
       shoulder(s)  or   a   valley   between   two   or   more maxima),
       appropriate  selection  of  analyte  spectra   and   background
       spectra  is  important.   Examination of extracted ion  current
       profiles  of appropriate  ions  can  aid  in the selection  of
       spectra,  and in qualitative  identification of compounds.  When
       analytes  coelute  (i.e.,  only  one chromatographic  peak  is
       apparent),  the identification  criteria  can be met, but  each
       analyte  spectrum  will  contain  extraneous  ions  contributed by
       the coeluting  compound.

       7.6.1.2      For samples containing components  not  associated
with the calibration  standards,  a  library search may be made  for the
purpose of tentative  identification.   The necessity  to perform  this
type of  identification  will  be  determined by  the type of analyses
being conducted. Guidelines for  making tentative identification  are:

       (1)    Relative intensities  of major ions  in  the  reference
             spectrum (ions > 10% of the most abundant ion)  should
             be present  in the sample spectrum.

       (2)    The relative  intensities of the major ions should agree
             within + 20%.   (Example:  For  an  ion with an  abundance
             of 50%  in  the  standard spectrum,  the corresponding
             sample ion  abundance  must be  between 30  and  70%).

       (3)    Molecular ions present in the reference  spectrum should
             be present  in the sample spectrum.

       (4)    Ions  present  in  the  sample  spectrum  but not  in  the
             reference  spectrum  should   be  reviewed  for possible
             background   contamination   or   presence  of   coeluting
             compounds.

       (5)    Ions  present  in  the reference spectrum but not  in  the
             sample  spectrum  should  be  reviewed  for  possible
             subtraction  from'  the  sample  spectrum  because   of
             background  contamination  or  coeluting  peaks.     Data
             system library reduction programs can sometimes  create
             these  discrepancies.

       Computer  generated  library  search  routines  should  not   use
normalization routines that would misrepresent the library  or  unknown
spectra when compared to each other.   Only after visual  comparison
of sample with the nearest library searches will the mass spectral
interpretation specialist assign a tentative identification.
                      8260A - 23                        Revision 1
                                                    September 1994

-------
7.6.2 Quantitative  analysis

      7.6.2.1      When   a   compound  has  been   identified,   the
quantitation  of  that  compound  will  be  based  on the  integrated
abundance  from   the   EICP   of  the  primary   characteristic  ion.
Quantitation will take place using the internal standard technique.
The  internal  standard  used  shall be the  one  nearest  the retention
time of that of a given analyte.

      7.6.2.2      When MS response is  linear  and passes through the
origin,  calculate the concentration of each identified  analyte in the
sample as follows:

      Water

                                (AJ(IS)
      concentration (jug/L)  = 	—	
                               (Ais)(RF)(V0)

where:

      Ax     =     Area of  characteristic  ion for compound  being
                   measured.
      Is     =     Amount of internal  standard injected  (ng).
      Ais     =     Area  of  characteristic  ion  for  the  internal
      	           standard.
      RF     =     Mean relative response factor for compound being
                   measured.
      V0     =     Volume  of   water  purged   (mL),   taking   into
                   consideration any dilutions made.

      Sediment/Soil  Sludge  (on  a  dry-weight  basis)  and  Waste
(normally on a wet-weight basis)

                                  (AJd.MV.)
      concentration (M9/kg)  =
                               (A,J(RF)(V,)(WS)(D)

where:

      Ax, Is,  Ais,  RF,  =  Same  as for water.
      Vt     =     Volume of total extract (p.1) (use 10,000 fj.1 or a
                   factor of this when dilutions are made).
      V|     =     Volume of extract added (/zL) for purging.
      Ws     =     Weight of sample extracted or purged  (g).
      D     =     % dry weight of sample/100,  or 1 for a wet-weight
                   basis.

      7.6.2.3      Where appl icable, an estimate of concentration for
noncalibrated components in the sample  should be made.  The formulae
given above  should  be used  with  the  following modifications:  The
areas Ax and Ais  should be from the total  ion  chromatograms, and the
RF for  the  compound  should be assumed to be 1.  The  concentration

                       8260A  -  24                         Revision 1
                                                     September 1994

-------
            obtained  should be  reported  indicating  (1)  that the value  is an
            estimate  and  (2)  which  internal  standard  was  used  to  determine
            concentration.     Use  the  nearest   internal   standard   free  of
            interferences.

                  7.6.2.4     Alternatively, the  regression  line fitted to the
            initial calibration  (Sec. 7.3.6.1) may be used for determination of
            analyte concentration.


8.0   QUALITY CONTROL

      8.1   Refer  to  Chapter One  and  Method 8000 for  general  quality control
procedures.

      8.2   Additional required  instrument QC is found in the  Sees. 7.3 and 7.4:

            8.2.1  The GC/MS system must be tuned to meet the BFB specifications.

            8.2.2  There  must be  an initial  calibration  of the GC/MS  system

            8.2.3  The GC/MS  system  must meet  the SPCC  criteria and the CCC
      criteria, each  12  hours.

      8.3   To  establish   the  ability  to   generate  acceptable  accuracy  and
precision,  the analyst must perform the following operations.

            8.3.1  A  quality control   (QC)  reference   sample  concentrate  is
      required containing each analyte  at a  concentration of  10 mg/L or less in
      methanol.  The  QC  reference sample concentrate may be prepared from pure
      standard materials or purchased  as certified  solutions.   If prepared by
      the laboratory, the  QC  reference sample  concentrate must  be  made  using
      stock standards prepared independently from those used for calibration.

            8.3.2  Prepare a QC reference sample to contain 20 /ug/L  or less of
      each  analyte by adding 200 /uL of QC reference sample concentrate to 100 mL
      of organic-free reagent water.

            8.3.3  Four 5-mL aliquots  of the well  mixed  QC reference  sample are
      analyzed according to the  method beginning in  Sec. 7.5.1.

            8.3.4  Calculate the  average recovery (x)  in /Ltg/L, and the standard
      deviation of the  recovery  (s)  in M9/U  for each  analyte  using  the four
      results.

            8.3.5  Tables  7  and 8  provide  single  laboratory  recovery  and
      precision data  obtained  for the method  analytes  from water.   Similar
      results from dosed water should be expected  by any experienced laboratory.
      Compare s and x   (Sec. 8.3.4) for each analyte to the  single  laboratory
      recovery and precision data.   Results are comparable  if  the  calculated
      standard deviation of the  recovery does  not exceed 2.6 times the single
      laboratory RSD or 20%, whichever is greater,  and the  mean recovery lies
      within the interval x ± 3s or  x  ± 30%, whichever  is greater.


                                  8260A -  25                        Revision 1
                                                                September 1994

-------
            NOTE:  The large  number  of  analytes  in Tables  7  and 8  present  a
                  substantial probability that one or  more  will  fail  at least
                  one of the acceptance criteria when  all  analytes  of a given
                  method are determined.

            8.3.6  When one or more of  the  analytes  tested are not comparable to
      the data  in Table 6  or  7,  the  analyst  must proceed according  to Sec.
      8.3.6.1 or 8.3.6.2.

                  8.3.6.1     Locate  and  correct the source of the problem and
            repeat the test  for  all  analytes  beginning  with  Sec.  8.3.2.

                  8.3.6.2     Beginning with  Sec.  8.3.2,  repeat  the test only
            for  those  analytes  that  are  not  comparable.   Repeated  failure,
            however,  will confirm a general problem  with the measurement system.
            If  this occurs,   locate  and  correct the source of  the  problem and
            repeat the test  for all  compounds  of interest  beginning with Sec.
            8.3.2.

      8.4   For  aqueous  and soil  matrices,  laboratory established  surrogate
control limits should be compared  with the control  limits  listed  in Table 8.

            8.4.1  If recovery is not within limits,  the  following procedures are
      required.

                  8.4.1.1     Check to  be sure that there are no errors in the
            calculations,  surrogate  solutions or internal standards.  If errors
            are  found,  recalculate the data  accordingly.

                  8.4.1.2     Check instrument performance.   If  an instrument
            performance  problem is  identified, correct the problem and re-analyze
            the  extract.

                  8.4,1.3     If no problem is found, re-extract and re-analyze
            the  sample.

                  8.4.1.4     If, upon  re-analysis,  the recovery is again not
            within limits, flag  the  data as  "estimated  concentration".

            8.4.2  At a minimum, each laboratory should update surrogate recovery
      limits on a matrix-by-matrix basis,  annually.


9.0   METHOD PERFORMANCE

      9.1   The  method   detection   limit  (MDL)  is  defined  as  the  minimum
concentration of   a  substance   that  can  be measured  and  reported with  99%
confidence that the value is above zero.  The MDL actually achieved in a given
analysis will vary depending on instrument sensitivity and matrix effects.

      9.2   This  method  has  been  tested  in  a single  laboratory  using spiked
water.  Using a wide-bore capillary column,  water was  spiked at concentrations
between 0.5  and  10 ng/l.   Single  laboratory accuracy and  precision  data are


                                  8260A -  26                        Revision  1
                                                                September 1994

-------
presented for the method analytes in Table  6.   Calculated MDLs are presented in
Table 1.
      9.3   The  method was tested  using water spiked  at  0.1  to 0.5  ng/l and
analyzed on a cryofocussed narrow-bore  column.  The accuracy and precision data
for these compounds are presented in Table 7.  MDL values were also calculated
from these data and are presented in Table 2.

      9.4   Direct  injection  has  been used for the analysis of waste motor oil
samples using a wide-bore column.   The accuracy  and  precision  data for these
compounds are presented in Table 10.


10.0  REFERENCES

1.    Methods for the Determination  of Organic Compounds  in  Finished Drinking
      Water and Raw  Source Water Method  524.2;  U.S.  Environmental  Protection
      Agency. Office  of  Research  Development,  Environmental   Monitoring  and
      Support Laboratory,  Cincinnati, OH 1986.

2.    U.S.  EPA  Contract  Laboratory Program,  Statement of  Work for Organic
      Analysis,  July 1985, Revision.

3.    Bellar, T.A.; J.J.  Lichtenberg. J. Amer. Water Works Assoc. 1974, 66(12),
      739-744.

4.    Bellar, T.A.;  J.J.   Lichtenberg.  "Semi -Automated Headspace Analysis  of
      Drinking  Waters  and  Industrial  Waters for Purgeable Volatile Organic
      Compounds";  in Van Hall, Ed.; Measurement of Organic Pollutants in Water
      and Wastewater, ASTM STP 686, pp 108-129,  1979.

5.    Budde, W.L.;  J.W.  Eichelberger.  "Performance Tests for the Evaluation of
      Computerized   Gas    Chromatography/Mass    Spectrometry   Equipment   and
      Laboratories";  U.S.  Environmental   Protection   Agency.    Environmental
      Monitoring and Support  Laboratory, Cincinnati, OH 45268, April 1980; EPA-
      600/4-79-020.

6.    Eichelberger,  J.W.; L.E.  Harris;  W.L.  Budde.   "Reference Compound  to
      Calibrate   Ion  Abundance   Measurement   in   Gas   Chromatography-Mass
      Spectrometry Systems";  Analytical Chemistry 1975, 47, 995-1000.

7.    Olynyk, P.; W.L.  Budde; J.W.  Eichelberger. "Method  Detection Limit for
      Methods 624 and 625"; Unpublished report,  October 1980.

8.    Non   Cryogenic   Temperatures   Program   and   Chromatogram,   Private
      Communications;  Myron  Stephenson   and  Frank   Allen,   EPA   Region  IV
      Laboratory,  Athens,  GA.

9.    Marsden, P.;  C.L. Helms, B.N. Colby.  "Analysis  of Volatiles  in Waste Oil";
      report for B. Lesnik, OSW/EPA under EPA contract 68-W9-001, 6/92.
                                  8260A - 27                        Revision 1
                                                                September 1994

-------
10.    Methods for  the Determination of  Organic Compounds  in  Drinking Water,
      Supplement II M6th_o_d_524.2;  U.S.  Environmental Protection Agency.  Office
      of Research and  Development,  Environmental Monitoring  Systems Laboratory,
      Cincinnati, OH  1992.
                                  8260A - 28                        Revision  1
                                                                September  1994

-------
                            TABLE  1.
CHROMATOGRAPHIC RETENTION TIMES AND METHOD DETECTION LIMITS (MDL)
  FOR VOLATILE ORGANIC COMPOUNDS ON WIDE-BORE CAPILLARY COLUMNS
ANALYTE
RETENTION TIME
(minutes)
Column la Column 2° Column 2'
Di chl orodi f 1 uoromethane
Chloromethane
Vinyl Chloride
Bromomethane
Chloroethane
Tri chl orofl uoromethane
Acrolein
lodomethane
Acetonitrile
Carbon disulfide
Ally! chloride
Methyl ene chloride
1,1-Dichloroethene
Acetone
trans-l,2-Dichloroethene
Acrylonitrile
1,1-Dichloroethane
Vinyl acetate
2,2-Dichloropropane
2-Butanone
cis-l,2-Dichloroethene
Propionitrile
Chloroform
Bromochl oromethane
Methacrylonitrile
1,1,1-Trichloroethane
Carbon tetrachloride
1,1-Dichloropropene
Benzene
1,2-Dichloroethane
Trichloroethene
1,2-Dichloropropane
Bromodi chl oromethane
Dibromomethane
Methyl methacrylate
1,4-Dioxane
2-Chloroethyl vinyl ether
4-Methyl -2-pentanone
trans-l,3-Dichloropropene
Toluene
cis-l,3-Dichloropropene
1 , 1 , 2 -Tri chl oroethane
1.35
1.49
1.56
2.19
2.21
2.42
3.19
3.56
4.11
4.11
4.11
4.40
4.57
4.57
4.57
5.00
6.14
6.43
8.10

8.25
8.51
9.01

9.19
10.18
11.02
--
11.50
12.09
14.03
14.51
15.39
15.43
15.50
16.17
--
17.32
17.47
18.29
19.38
19.59
0.70
0.73
0.79
0.96
1.02
1.19





2.06
1.57

2.36

2.93

3.80

3.90

4.80
4.38

4.84
5.26
5.29
5.67
5.83
7.27
7.66
8.49
7.93




--
10.00
--
11.05
3.13
3.40
3.93
4.80

6.20





9.27
7.83

9.90

10.80

11.87

11.93

12.60
12.37

12.83
13.17
13.10
13.50
13.63
14.80
15.20
15.80
15.43




16.70
17.40
17.90
18.30
MDLd
(M9/L)
0.10
0.13
0.17
0.11
0.10
0.08





0.03
0.12

0.06

0.04

0.35

0.12

0.03
0.04

0.08
0.21
0.10
0.04
0.06
0.19
0.04
0.08
0.24





0.11
--
0.10
                           8260A  -  29
    Revision 1
September 1994

-------
TABLE 1.
(Continued)
ANALYTE


Ethyl methacrylate
2-Hexanone
Tetrachloroethene
1,3-Dichloropropane
Di bromochl oromethane
1,2-Dibromoethane
1-Chlorohexane
Chlorobenzene
1,1,1 , 2-Tetrachl oroethane
Ethyl benzene
p-Xylene
m-Xylene
o-Xylene
Styrene
Bromoform
Isopropyl benzene (Cumene)
cis-l,4-Dichloro-2-butene
1 , 1 , 2, 2-Tetrachl oroethane
Bromobenzene
1,2,3-Trichloropropane
n-Propyl benzene
2-Chlorotoluene
trans-l,4-Oichloro-2-butene
1, 3, 5 -Tri methyl benzene
4-Chlorotoluene
Pentachl oroethane
1, 2, 4-Trimethyl benzene
sec-Butyl benzene
tert-Butyl benzene
p- Isopropyl toluene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
Benzyl chloride
n-Butyl benzene
1,2-Dichlorobenzene
1 ,2-Dibromo-3-chloropropane
1 ,2,4-Trichlorobenzene
Hexachl orobutadi ene
Naphthalene
1, 2, 3-Tri chlorobenzene
RETENTION TIME

Column 1"
20.01
20.30
20.26
20.51
21.19
21.52
--
23.17
23.36
23.38
23.54
23.54
25.16
25.30
26.23
26.37
27.12
27.29
27.46
27.55
27.58
28.19
28.26
28.31
28.33
29.41
29.47
30.25
30.59
30.59
30.56
31.22
32.00
32.23
32.31
35.30
38.19
38.57
39.05
40.01
(minutes)
Column 2"


11.15
11.31
11.85
11.83
13.29
13.01
13.33
13.39
13.69
13.68
14.52
14.60
14.88
15.46

16.35
15.86
16.23
16.41
16.42

16.90
16.72

17.70
18.09
17.57
18.52
18.14
18.39

19.49
19.17
21.08
23.08
23.68
23.52
24.18

Column 2'c


18.60
18.70
19.20
19.40
--
20.67
20.87
21.00
21.30
21.37
22.27
22.40
22.77
23.30

24.07
24.00
24.13
24.33
24.53

24.83
24.77

31.50
26.13
26.60
26.50
26.37
26.60

27.32
27.43
--
31.50
32.07
32.20
32.97
MDLd
(M9/L)



0.14
0.04
0.05
0.06
0.05
0.04
0.05
0.06
0.13
0.05
0.11
0.04
0.12
0.15

0.04
0.03
0.32
0.04
0.04

0.05
0.06

0.13
0.13
0.14
0.12
0.12
0.03

0.11
0.03
0.26
0.04
0.11
0.04
0.03
8260A - 30
    Revision 1
September 1994

-------
                                   TABLE 1.
                                  (Continued)
ANALYTE                                       RETENTION  TIME               MDLd
                                    	(minutes)	  (M9/U
                                    Column  1"Column  2bColumn  2/c
INTERNAL STANDARDS/SURROGATES
1,4-Difluorobenzene
Chlorobenzene-d5
l,4-Dichlorobenzene-d4
4-Bromofl uorobenzene
1 , 2-Dichl orobenzene-d4
Dichloroethane-d4
Di bromof 1 uoromethane
Toluene-d8
Pentaf 1 uorobenzene
Fl uorobenzene
13.26
23.10
31.16
27.83
32.30
12.08

18.27
--
13.00



15.71 23.63
19.08 27.25




6.27 14.06
8  Column 1 - 60 meter x 0.75 mm ID VOCOL capillary.  Hold at 10°C for 8  minutes,
   then program  to 180°C at 4%nn.

b  Column 2-30 meter x 0.53 mm  ID DB-624 wide-bore capillary  using cryogenic
   oven.  Hold at 10°C for 5 minutes,  then program to 160°C at 6°/min.

0  Column 2' - 30 meter  x 0.53 mm  ID DB-624 wide-bore capillary,  cooling GC oven
   to ambient  temperatures.   Hold  at  10°C  for 6 minutes,  program  to  70°C  at
   10°/min,  program to 120°C  at  5°/min,  then  program to  180°C at 8°/min.

d  MDL based on  a 25 mL sample volume.
                                  8260A - 31                        Revision  1
                                                                September  1994

-------
                            TABLE  2.
CHROMATOGRAPHIC RETENTION TIMES AND METHOD DETECTION LIMITS (MDL)
 FOR VOLATILE ORGANIC COMPOUNDS ON NARROW-BORE CAPILLARY COLUMNS
ANALYTE
                             RETENTION TIME
                                (minutes)
                                Column 3"
MDLb
Di chl orodi f 1 uoromethane
Chl oromethane
Vinyl chloride
Bromomethane
Chloroethane
Trichlorofl uoromethane
1,1-Dichloroethene
Methylene chloride
trans-l,2-Dichloroethene
1,1-Dichloroethane
cis-l,2-Dichloroethene
2,2-Dichloropropane
Chloroform
Bromochl oromethane
1,1,1-Tri chl oroethane
1,2-Dichloroethane
1,1-Dichloropropene
Carbon tetrachloride
Benzene
1,2-Dichloropropane
Trichloroethene
Dibromomethane
Bromodi chl oromethane
Toluene
1 , 1 , 2-Tri chl oroethane
1,3-Dichloropropane
Di bromochl oromethane
Tetrachloroethene
1,2-Dibromoethane
Chlorobenzene
1,1,1, 2-Tetrachl oroethane
Ethyl benzene
p-Xylene
m-Xylene
Bromoform
o-Xylene
Styrene
1,1,2 , 2-Tetrachl oroethane
1,2,3-Trichloropropane
Isopropyl benzene
0.88
0.97
1.04
1.29
1.45
1.77
2.33
2.66
3.54
4.03
5.07
5.31
5.55
5.63
6.76
7.00
7.16
7.41
7.41
8.94
9.02
9.09
9.34
11.51
11.99
12.48
12.80
13.20
13.60
14.33
14.73
14.73
15.30
15.30
15.70
15.78
15.78
15.78
16.26
16.42
0.11
0.05
0.04
0.06
0.02
0.07
0.05
0.09
0.03
0.03
0.06
0.08
0.04
0.09
0.04
0.02
0.12
0.02
0.03
0.02
0.02
0.01
0.03
0.08
0.08
0.08
0.07
0.05
0.10
0.03
0.07
0.03
0.06
0.03
0.20
0.06
0.27
0.20
0.09
0.10
                           8260A  - 32
       Revision 1
   September 1994

-------
                                   TABLE 2.
                                  (Continued)
ANALYTE
                                    RETENTION TIME
                                      (minutes)
                                      Column 3a
 MDLb
(M9/L)
Bromobenzene
2-Chlorotoluene
n-Propyl benzene
4-Chlorotoluene
1,3, 5-Tri methyl benzene
tert-Butyl benzene
1 , 2 , 4-Trimethyl benzene
sec-Butyl benzene
1,3-Dichlorobenzene
p-Isopropyltoluene
1,4-Dichlorobenzene
1,2-Dichlorobenzene
n-Butyl benzene
1 ,2-Dibromo-3-chloropropane
1,2,4-Trichlorobenzene
Naphthalene
Hexachlorobutadiene
1,2,3-Trichlorobenzene
16.42
16.74
16.82
16.82
16.99
17.31
17.31
17.47
17.47
17.63
17.63
17.79
17.95
18.03
18.84
19.07
19.24
19.24
0.11
0.08
0.10
0.06
0.06
0.33
0.09
0.12
0.05
0.26
0.04
0.05
0.10
0.50
0.20
0.10
0.10
0.14
a  Column 3 - 30 meter x 0.32 mm ID DB-5 capillary with 1

b  HDL based on a 25 ml sample volume.
                                                             film thickness,
                                  8260A - 33
                                                                    Revision 1
                                                                September 1994

-------
                                 TABLE 3.
           ESTIMATED QUANTITATION LIMITS  FOR VOLATILE ANALYTES8
                                     Estimated Quantitation Limits
                                       (All Analytes in Table 1)
                                  Ground water         Low Soil/Sediment6
    Purging 5 mL of water              5

    Purging 25 mL of water             1

    Soil/Sediment
   Estimated Quantitation Limit (EQL) - The lowest concentration that can be
   reliably  achieved within specified limits of precision  and accuracy during
   routine   laboratory operating conditions.  The  EQL  is generally  5  to 10
   times the MDL.  However, it may be nominally chosen within these guidelines
   to simplify data   reporting. For many analytes the EQL  is selected  from the
   lowest non-zero standard in the calibration curve. Sample EQLs are highly
   matrix-dependent. The EQLs  listed herein are provided for guidance and may
   not always be achievable.

   EQLs listed for soil/sediment are based  on wet  weight.  Normally data are
   reported on a dry weight basis;  therefore,  EQLs  will  be  higher, based on
   the percent dry weight in  each sample.
             Other Matrices                      Factor0
             Water miscible liquid waste             50
             High-concentration soil  and sludge     125
             Non-water miscible waste               500


CEQL  =   [EQL for low soil/sediment (see  Table 3)] X [Factor], For non-aqueous
        samples, the factor is on a wet-weight basis.
                                8260A - 34                        Revision 1
                                                              September 1994

-------
                             TABLE 4.
    BFB MASS  -  INTENSITY  SPECIFICATIONS  (4-BROMOFLUOROBENZENE)"


Mass              Intensity Required (relative abundance)


 50               15 to 40% of mass 95
 75               30 to 60% of mass 95
 95               base peak,  100% relative abundance
 96               5 to 9% of mass 95
173               less than 2% of mass 174
174               greater than 50% of mass 95
175               5 to 9% of mass 174
176               greater than 95% but less than 101% of mass 174
177               5 to 9% of mass 176
     Alternate tuning criteria may  be  used (e.g. CLP, Method  524.2,  or
     manufacturers'  instructions), provided that method performance is not
     adversely affected.
                            8260A - 35                        Revision 1
                                                          September 1994

-------
                             TABLE 5.
    CHARACTERISTIC  MASSES  (M/Z)  FOR  PURGEABLE  ORGANIC  COMPOUNDS
Analyte
  Primary
Characteristic
    Ion
  Secondary
Characteristic
    Ion(s)
Acetone
Acetonitrile
Acrolein
Acrylonitrile
Allyl alcohol
Ally! chloride
Benzene
Benzyl chloride
Bromoacetone
Bromobenzene
Bromochl oromethane
Bromodi chl oromethane
Bromoform
Bromomethane
iso-Butanol
n-Butanol
2-Butanone
n-Butyl benzene
sec-Butyl benzene
tert-Butyl benzene
Carbon disulfide
Carbon tetrachloride
Chloral hydrate
Chloroacetonitrile
Chlorobenzene
1-Chlorobutane
Chl orodi bromomethane
Chloroethane
2-Chloroethanol
bis-(Z-chloroethyl) sulfide
2-Chloroethyl vinyl ether
Chloroform
Chl oromethane
Chloroprene
3-Chloropropionitrile
2-Chlorotoluene
4-Chlorotoluene
l,2-Dibromo-3-chloropropane
Di bromochl oromethane
1,2-Dibromoethane
Di bromomethane
1,2-Dichlorobenzene
l,2-Dichlorobenzene-d4
58
41
56
53
57
76
78
91
136
156
128
83
173
94
74
56
72
91
105
119
76
117
82
48
112
56
129
64(49*)
49
109
63
83
50(49*)
53
54
91
91
75
129
107
93
146
152
43
41, 40, 39
55, 58
52, 51
57, 58, 39
76, 41, 39, 78
-
91, 126, 65, 128
43, 136, 138, 93, 95
77, 158
49, 130
85, 127
175, 254
96
43
41
43, 72
92, 134
134
91, 134
78
119
44, 84, 86, 111
75
77, 114
49
208, 206
66(51*)
49, 44, 43, 51, 80
111, 158, 160
65, 106
85
52(51*)
53, 88, 90, 51
54, 49, 89, 91
126
126
155, 157
127
109, 188
95, 174
111, 148
115, 150
                            8260A - 36
                                Revision 1
                            September 1994

-------
                        TABLE  5.(continued)
Analyte
  Primary
Characteristic
    Ion
  Secondary
Characteristic
    Ion(s)
1,3-Dichlorobenzene
1,4-Dichlorobenzene
cis-l,4-Dichloro-2-butene
trans-l,4-Dichloro-2-butene
Dichlorodifluoromethane
1,1-Dichloroethane
1,2-Dichloroethane
1,1-Dichloroethene
cis-l,2-Dichloroethene
trans-l,2-Dichloroethene
1,2-Dichloropropane
1,3-Dichloropropane
2,2-Dichloropropane
l,3-Dichloro-2-propanol
1 ,1-Dichloropropene
cis-l,3-Dichloropropene
trans-l,3-Dichloropropene
1,2,3,4-Diepoxybutane
Diethyl ether
1,4-Dioxane
Epichlorohydrin
Ethanol
Ethyl acetate
Ethyl benzene
Ethylene oxide
Ethyl methacrylate
Hexachlorobutadiene
Hexachloroethane
2-Hexanone
2-Hydroxypropionitrile
lodomethane
Isobutyl alcohol
Isopropyl benzene
p-Isopropyltoluene
Malononitrile
Methacrylonitrile
Methyl acrylate
Methyl -t-butyl ether
Methylene chloride
Methyl ethyl ketone
Methyl iodide
Methyl methacrylate
4-Methyl -2-pentanone
Naphthalene
Nitrobenzene
146
146
75
53
85
63
62
96
96
96
63
76
77
79
75
75
75
55
74
88
57
31
88
91
44
69
225
201
43
44
142
43
105
119
66 '
41
55
73
84
72
142
69
100
128
123
111, 148
111, 148
75, 53, 77, 124,
88, 75
87
65, 83
98
61, 63
61, 98
61, 98
112
78
97
79, 43, 81, 49
110, 77
77, 39
77, 39
55, 57, 56
45, 59
88, 58, 43, 57
57, 49, 62, 51
45, 27, 46
43, 45, 61
106
44, 43, 42
69, 41, 99, 86,
223, 227
166, 199, 203
58, 57, 100
44, 43, 42, 53
127, 141
43, 41, 42, 74
120
134, 91
66, 39, 65, 38
41, 67, 39, 52,
85
57
86, 49
43
142, 127, 141
69, 41, 100, 39
43, 58, 85
-
51, 77


89






















114









66









                             8260A  -  37
                                Revision  1
                            September  1994

-------
                        TABLE  5.(continued)
Analyte
  Primary
Characteristic
    Ion
  Secondary
Characteristic
    Ion(s)
2-Nitropropane
2-Picoline
Pentachloroethane
Propargyl alcohol
6-Propiolactone
Propionitrile (ethyl cyanide)
n-Propylamine
n-Propyl benzene
Pyridine
Styrene
1,2,3-Trichlorobenzene
1,2,4-Trichlorobenzene
1,1,1 , 2-Tetrachloroethane
1,1,2 , 2-Tetrachl oroethane
Tetrachloroethene
Toluene
1,1,1-Trichloroethane
1, 1, 2 -Trichl oroethane
Trichloroethene
Tri chl orof 1 uoromethane
1,2,3-Trichloropropane
1, 2, 4-Trimethyl benzene
1,3, 5- Trimethyl benzene
Vinyl acetate
Vinyl chloride
o-Xylene
m-Xylene
p-Xylene
INTERNAL STANDARDS/SURROGATES
1,4-Difluorobenzene
Chlorobenzene-d5
l,4-Dichlorobenzene-d4
4-Bromofl uorobenzene
Di bromof 1 uoromethane
Dichloroethane-d4
Toluene-d8
Pentaf 1 uorobenzene
Fl uorobenzene
46
93
167
55
42
54
59
91
79
104
180
180
131
83
164
92
97
83
95
151
75
105
105
43
62
106
106
106

114
117
152
95
113'
102
98
168
96
.
93,
167,
55,
42,
54,
59,
120
52
78
182,
182,
133,
131,
129,
91
99,
97,
97,
101,
77
120
120
86
64
91
91
91



115,
174,




77

66, 92, 78
130, 132, 165, 169
39, 38, 53
43, 44
52, 55, 40
41, 39



145
145
119
85
131, 166

61
85
130, 132
153











150
176





* - characteristic ion for an  ion trap mass spectrometer (to be used when
ion-molecule reactions are observed)
                            8260A - 38
                                Revision 1
                            September 1994

-------
                         TABLE 6.
SINGLE LABORATORY ACCURACY AND PRECISION DATA FOR VOLATILE
    ORGANIC COMPOUNDS IN WATER DETERMINED WITH A WIDE-
                  BORE CAPILLARY COLUMN
Analyte
Benzene
Bromobenzene
Bromochl oromethane
Bromodi chl oromethane
Bromoform
Bromomethane
n-Butyl benzene
sec-Butyl benzene
tert-Butyl benzene
Carbon tetrachloride
Chlorobenzene
Chloroethane
Chloroform
Chl oromethane
2-Chlorotoluene
4-Chlorotoluene
1 , 2-Dibromo-3-Chl oropropane
Di bromochl oromethane
1,2-Dibromoethane
Dibromomethane
1 , 2-Di chl orobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
Di chl orodi f 1 uoromethane
1,1-Dichlorobenzene
1,2-Dichlorobenzene
1,1-Dichloroethene
cis-1, 2-Di chl oroethene
trans -1, 2-Di chl oroethene
1,2-Dichloropropane
1,3-Dichloropropane
2,2-Dichloropropane
1,1-Dichloropropene
Ethyl benzene
Hexachl orobutadi ene
Isopropyl benzene
p-Isopropyl toluene
Methylene chloride
Naphthalene
n-Propyl benzene
Styrene
Cone. Number
Range, of Recovery8
pig/L Samples %
0.1
0.1
0.5
0.1
0.5
0.5
0.5
0.5
0.5
0.5
0.1
0.5
0.5
0.5
0.1
0.1
0.5
0.1
0.5
0.5
0.1
0.5
0.2
0.5
0.5
0.1
0.1
0.5
0.1
0.1
0.1
0.5
0.5
0.1
C.5
0.5
0.1
0.1
0.1
0.1
0.1
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 20
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
-100
- 10
-100
31
30
24
30
18
18
18
16
18
24
31
24
24
23
31
31
24
31
24
24
31
24
31
18
24
31
34
18
30
30
31
12
18
31
18
16
23
30
31
31
39
97
100
90
95
101
95
100
100
102
84
98
89
90
93
90
99
83
92
102
100
93
99
103
90
96
95
94
101
93
97
96
86
98
99
100
101
99
95
104
100
102
Standard
Deviation Percent
of Recovery13 RSD
6.5
5.5
5.7
5.7
6.4
7.8
7.6
7.6
7.4
7.4
5.8
8.0
5.5
8.3
5.6
8.2
16.6
6.5
4.0
5.6
5.8
6.8
6.6
6.9
5.1
5.1
6.3
6.7
5.2
5.9
5.7
14.6
8.7
8.4
6.8
7.7
6.7
5.0
8.6
5.8
7.3
5.7
5.5
6.4
6.1
6.3
8.2
7.6
7.6
7.3
8.8
5.9
9.0
6.1
8.9
6.2
8.3
19.9
7.0
3.9
5.6
6.2
6.9
6.4
7.7
5.3
5.4
6.7
6.7
5.6
6.1
6.0
16.9
8.9
8.6
6.8
7.6
6.7
5.3
8.2
5.8
7.2
                       8260A  - 39
    Revision 1
September 1994

-------
                                   TABLE 6.
                                  (Continued)


Analyte
Cone.
Range,
M9A
Number
of Recovery8
Samples %
Standard
Deviation Percent
of Recovery15 RSD
1,1,1,2-Tetrachloroethane   0.5 - 10      24     90
1,1,2,2-Tetrachloroethane   0.1 - 10      30     91
Tetrachloroethene           0.5 - 10      24     89
Toluene                     0.5 - 10      18    102
1,2,3-Trichlorobenzene      0.5 - 10      18    109
1,2,4-Trichlorobenzene      0.5 - 10      18    108
1,1,1-Trichloroethane       0.5 - 10      18     98
1,1,2-Trichloroethane       0.5 - 10      18    104
Trichloroethene             0.5 - 10      24     90
Trichlorofluoromethane      0.5 - 10      24     89
1,2,3-Trichloropropane      0.5 - 10      16    108
1,2,4-Trimethylbenzene      0.5-10      18     99
1,3,5-Trimethylbenzene      0.5 - 10      23     92
Vinyl chloride              0.5 - 10      18     98
o-Xylene                    0.1 - 31      18    103
m-Xylene                    0.1 - 10      31     97
p-Xylene                    0.5 - 10      18    104
  6.1
 5.7
 6.0
 8.1
 9.4
 9.0
 7.9
 7.6
 6.5
 7.2
15.6
 8.0
 6.8
 6.5
 7.4
 6,3
 8.0
 6.8
 6.3
 6.8
 8.0
 8.6
 8.3
 8.1
 7.3
 7.3
 8.1
14.4
 8.1
 7.4
 6.7
 7.2
 6.5
 7.7
8  Recoveries were calculated using  internal standard method. Internal standard
   was fluorobenzene.

b  Standard deviation was calculated by  pooling data from three concentrations.
                                  8260A - 40
         Revision 1
     September 1994

-------
                    TABLE 7.
SINGLE LABORATORY ACCURACY AND PRECISION DATA FOR
 VOLATILE ORGANIC COMPOUNDS  IN WATER DETERMINED
       WITH A NARROW-BORE CAPILLARY COLUMN
Analyte
Benzene
Bromobenzene
Bromochl oromethane
Bromodi chl oromethane
Bromoform
Bromomethane
n-Butyl benzene
sec-Butyl benzene
tert-Butyl benzene
Carbon tetrachloride
Chlorobenzene
Chloroethane
Chloroform
Chloromethane
2-Chlorotoluene
4-Chlorotoluene
1 ,2-Dibromo-3-chl oropropane
Di bromochl oromethane
1,2-Dibromoethane
Dibromomethane
1, 2 -Di chlorobenzene
1 ,3-Dichlorobenzene
1,4-Dichlorobenzene
Di chl orodi f 1 uoromethane
1,1-Dichloroethane
1,2-Dichloroethane
1,1-Dichloroethene
cis-l,2-Dichloroethene
trans-l,2-Dichloroethene
1,2-Dichloropropane
1,3-Dichloropropane
2,2-Dichloropropane
1,1-Dichloropropene
Ethyl benzene
Hexachlorobutadiene
I sopropyl benzene
p-Isopropyltoluene
Methylene chloride
Naphthalene
n-Propyl benzene
Cone.
M9/L
0.1
0.5
0.5
0.1
0.5
0.5
0.5
0.5
0.5
0.1
0.1
0.1
0.1
0.5
0.5
0.5
0.5
0.1
0.5
0.5
0.1
0.1
0.1
0.1
0.5
0.1
0.1
0.1
0.1
0.5
0.5
0.5
0.5
0.5
0.5
0.5
0.5
0.5
0.5
0.5
Number
of
Samples
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
Recovery8
%
99
97
97
100
101
99
94
110
110
108
91
100
105
101
99
96
92
99
97
93
97
101
106
99
98
100
95
100
98
96
99
99
102
99
100
102
113
97
98
99
Standard
Deviation
of Recovery
6.2
7.4
5.8
4.6
5.4
7.1
6.0
7.1
2.5
6,8
5.8
5.8
3.2
4.7
4.6
7.0
10.0
5.6
5.6
5.6
3.5
6.0
6.5
8.8
6.2
6.3
9.0
3.7
7.2
6.0
5.8
4.9
7.4
5.2
6.7
6.4
13.0
13.0
7.2
6.6
Percent
RSD
6.3
7.6
6.0
4.6
5.3
7.2
6.4
6.5
2.3
6.3
6.4
5.8
3.0
4.7
4.6
7.3
10.9
5.7
5.8
6.0
3.6
5.9
6.1
8.9
6.3
6.3
9.5
3.7
7.3
6.3
5.9
4.9
7.3
5.3
6.7
6.3
11.5
13.4
7.3
6.7
                   8260A - 41
    Revision 1
September 1994

-------
TABLE 7.
(Continued)


Analyte
Styrene
1,1,1 , 2-Tetrachl oroethane
1,1,2 , 2-Tetrachl oroethane
Tetrachloroethene
Toluene
1 , 2 , 3-Tri chl orobenzene
1 , 2 , 4-Tr i chl orobenzene
1,1,1-Tri chl oroethane
1,1,2-Trichloroethane
Trichloroethene
Tr i chl orof 1 uoromethane
1,2, 3-Tri chl oropropane
1 , 2, 4-Trimethyl benzene
1 ,3 , 5-Tri methyl benzene
Vinyl chloride
o-Xylene
m-Xylene
p-Xylene

Cone.
yug/L
0.5
0.5
0.5
0.1
0.5
0.5
0.5
0.5
0.5
0.1
0.1
0.5
0.5
0.5
0.1
0.5
0.5
0.5
Number
of
Samples
7
7
7
7
7
7
7
7
7
7
7
7
7
1
1
1
1
7

Recovery"
%
96
100
100
96
100
102
91
100
102
104
97
96
96
101
104
106
106
97
Standard
Deviation
of Recovery
19.0
4.7
12.0
5.0
5.9
8.9
16.0
4.0
4.9
2.0
4.6
6.5
6.5
4.2
0.2
7.5
4.6
6.1

Percent
RSD
19.8
4.7
12.0
5.2
5.9
8.7
17.6
4.0
4.8
1.9
4.7
6.8
6.8
4.2
0.2
7.1
4.3
6.3
Recoveries were calculated using  internal  standard method.  Internal standard
was fluorobenzene.
                               8260A - 42
    Revision 1
September 1994

-------
                                   TABLE 8.
      SURROGATE SPIKE RECOVERY LIMITS FOR WATER AND SOIL/SEDIMENT SAMPLES


Surrogate Compound
4-Bromofl uorobenzene8
Di bromof 1 uoromethane"
Toluene-d8a
Dichloroethane-d4a
Percent
Low/High
Water
86-115
86-118
88-110
80-120
Recovery
Low/High
Soil/Sediment
74-121
80-120
81-117
80-120
   Single laboratory data, for guidance only.
                                   TABLE 9.
                 QUANTITY OF EXTRACT REQUIRED FOR ANALYSIS OF
                          HIGH-CONCENTRATION SAMPLES
Approximate                                     Volume of
Concentration Range                             Extract8


   500 -  10,000 /xg/kg                          100 ML
 1,000 -  20,000 MQ/kg                           50 /iL
 5,000 - 100,000 jig/kg                           10 ML
25,000 - 500,000 /ug/kg                          100 /xL of  1/50 dilution15


Calculate appropriate dilution factor for concentrations exceeding this table.

a     The volume of solvent added to 5 mL of water being purged should be kept
      constant.  Therefore, add to the 5 mL syringe whatever volume of solvent
      is necessary to maintain a volume of 100 /xL added to the syringe.

b     Dilute  an aliquot  of  the  solvent  extract  and  then  take  100 /j,i  for
      analysis.
                                  8260A - 43                        Revision 1
                                                                September 1994

-------
                             TABLE 10
           DIRECT INJECTION  ANALYSIS  OF NEW OIL AT 5 PPM

Compound
Acetone
Benzene
n-Butanol*,**
iso-Butanol*,**
Carbon tetrachloride
Carbon disulfide**
Chlorobenzene
Chloroform
1,4-Dichlorobenzene
1,2-Dichloroethane
1,1-Dichloroethene
Di ethyl ether
Ethyl acetate
Ethyl benzene
Hexachloroethane
Methylene chloride
Methyl ethyl ketone
MIBK
Nitrobenzene
Pyridine
Tetrachloroethene

Recovery (%)
91
86
107
95
86
53
81
84
98
101
97
76
113
83
71
98
79
93
89
31
82
Trichlorofluoromethane 76
l,l,2-Cl3F3ethane
Toluene
Trichloroethene
Vinyl chloride
o-Xylene
m/p-Xylene
* Alternate mass
** T<^ nnant i t at i nr
69
73
66
63
83
84
employed
^

%RSD
14.8
21.3
27.8
19.5
44.7
22.3
29.3
29.3
24.9
23.1
45.3
24.3
27.4
30.1
30.3
45.3
24.6
31.4
30.3
35.9
27.1
27.6
29.2
21.9
28.0
35.2
29.5
29.5

Blank
(ppm)
1.9
0.1
0.5
0.9
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.2
0.0
0.0
0.4
0.0
0.0
0.0
0.0
0.0
0.0
0.6
0.0
0.0
0.4
0.6

Spike
(ppm)
5.0
0.5
5.0
5.0
0.5
5.0
5.0
6.0
7.5
0.5
0.7
5.0
5.0
5.0
3.0
5.0
5.0
5.0
2.0
5.0
0.7
5.0
5.0
5.0
0.5
0.2
5.0
10.0

Data are taken from Reference 9.
                            8260A - 44
    Revision 1
September 1994

-------
                            FIGURE 1.
                         PURGING DEVICE
 FOAM TIM*
EXrr IM m. 0.0.
                        GUT 1M M OJO.
                        14 MM O.O.
                        INLET 1M Ht. 0.0.
10 MM GLASS FWT
MCOUM
      INLET

2-WAY SYMNOE VALVE

17 CM 20 GAUGE SYftMQE NEEDLE

• MM O.O. RUeKA SEPTUM

INLET 1M IN. O.O.
                                                      1DQ IN 00
                                                   /~ STAINLESS STEEL
                                                     13X
                                                     MOLECULAR SIEVE
                                                     PUBOE GAS FH.TCT
                                                       PUMQEGAS
                                                       fl.CV» CXXTWX
                          8260A -  45
                                   Revision  1
                               September  1994

-------
                           FIGURE 2.
TRAP  PACKING AND CONSTRUCTION TO  INCLUDE DESORB CAPABILITY
      PACKING DETAIL

          i

         ZjT SMMOLAS
CONSTRUCTION DETAIL
          77 CM SnXA GEL
          IS CM TOUX QC
           •- i CM 3H OV-1
             5 MM OLAM »VOCX
                                               TH0MOCOUFIB
                                               COMTNOU9
                         8260A - 46
                          Revision  1
                      September  1994

-------
                            FIGURE 3.
          SCHEMATIC OF  PURGE-AND-TRAP DEVICE - PURGE MODE
CARRIER GAS
FLOW CONTROL
PRESSURE
REGULATOR
PURGE OAS
FLOW CONTROL
13X MOLECULAR
SIEVE FILTER
UCMO INJECTION PORTS

      COLUMN OVEN
                                    L/in/v
                                                CONFIRMATORY COLUMN
                                               TO DETECTOR
                                               > ANALYTICAL COLUMN
                              OPTIONAL 4^>ORT COLUMN
                              SELECTION VALVE
                                       TRAP INLET
                                      TRAP
                                PURGING
                                DEVICE
             NOTE
             ALL UN£S BETWEEN TRAP
             ANO OC SHOULD BE HEATED
             TO KTC
                            8260A -  47
                            Revision  1
                        September  1994

-------
                            FIGURE  4.
         SCHEMATIC OF PURGE-AND-TRAP DEVICE - DESORB MODE
CARRKRGAS
FLOW CONTROL
PRESSURE
REGULATOR
LJQUK) INJECTION PORTS
   r- COLUMN OVEN
                               OPTIONAL **ORT COLUMN
                               SELECTION VALVE
               CONFIRMATORY COLUMN

              TO DETECTOR
                                                 ANALYTICAL COLUMN
                                        TRAP INLET
PURGE GAS
FLOW CONTROL
13X MOLECULAR
SIEVE FILTER
                                 PURGING
                                 DEVCE
              NOTE
              ALL UNES BETWEEN TRAP
              AND OC SHOULD BE HEATED
              TOCTC.
                            8260A  -  48
                             Revision 1
                         September 1994

-------
                         FIGURE 5.

                 GAS CHROMATOGRAM OF VOLATILE ORGANICS
       408
            I 1C
               999
  1299
                                           2008
                                                   24OO
COLUMNi £0 METER * O.73 MM l.O. VOCOL CAPILLARY




PROGRAM* 10 C FOR 9 MIN., THEN 6 /HIN TO ISO C
  UJ
  O
  •_•
  oc
  O


  *

  J

  z
      2

      *

   u  •-


   5 *B s
   T »-• Of fc
   H *0 |
   u o J *

   g J?S u
   O ^ *^ «p



   " U «l A

   « s!-2


   i f I"
     u. i
^
   Id
   O
   oe
   a
&
 I  I il ~         I      I  I %  111

.M>M.A.4'^"*.w.ff
  2    4    6    8   18   12   14    16
                            Z it)

                            Ss

                            z d
                            a T
                            o P
                            or T
                            o a


                            i»
                            •-«
                            oe
                            H

                            I
                            «

                            N


                            -I
                            UJ


                            5
                            N


                            III
                            n

                            o
                            K

                            o

                            _i


                            u


                            TL

                            I


                            I

                            n


                            
-------
                                              FIGURE  6.
                                GAS CHROMATOGRAM OF VOLATILE ORGANICS
Column Z - 30m  long x O.SJnw 10 06-624
           •woe-bore col urn
PROGRAMi  10 C FOB 3 HIM.,
       THEN 6 /MIN TO  ISO C
                                         ftCTIMttOM TIMI. MIN.
                                              8260A - 50
    Revision 1
September 1994

-------
                                             FIGURE  7.
                            GAS  CHROMATOGRAM  OF VOLATILE  ORGANICS
 RIC
 04.-'1-9'87  9:26:00
                                                             SuJJS  125 TO 900
                                       Ctt.1: 40MS04238? «3
       COItlS.: F4000.40-160X8.12,F4,3mPURGE.TEIKIl.&EL.06624.SHEEP35,18PSI
            : G   1,1200  LftBEL: tl  8, 4.6  GUrtl: ft  8- 1.0 J  0  EASE:  U 28,  3
100.8-1
 F.1C
thane
oride
thane
thar.e


DC Et
lch
cme
roe
clilo
viny
brcr
chlo
C13

1
      ^
      i—1
      U




AAAJ'J
                                                   ij
                                                   ro
                                             C

                                             5H
                                    X3   K  *~*
                                    Wo ->  -I a«P
                                       (NO  o air<
                                        fi
                                        0) a
                                          *


                                                     (I

                                                    li
                                                     VL-
                                                       I
~°§   <
  ri&   c
  rH5?   fl)

       I-
       o
  667
       643
                                                                          705o
                                                                        s
                                                                        in
                                                                      e.
                                                                      a
                                                                                       u
                                                                                       »
                                                                                                    2641W.
             I'WJ
                                400
                                6:40
                                                                         11:40
                                                                                     6C>0
                                      iit>0  CJ'H(I
                                     IS:00 HMf
                                            8260A  - 51
                                                                                                   Revision  1
                                                                                              September  1994

-------
                                                                 FIGURE 8.
                                                  GAS  CHROMATOGRAM OF  TEST MIXTURE
(OT.I
                                   U
                                9
                                H
Q
O
                                                     KAMI
                                                      UN
                                                        0.5  g/L PER COMPOUND

                                                        1.   1,1-DICHLOROETHYLENE
                                                        2.   METHYLENE CHLORIDE
                                                        3.   TRANS-1.2-DICHLOROETHYLENE
                                                        4.   1,1 DICHLOROETHANE
                                                        5.   ISOPROPYLETHER
                                                        6.   CHLOROFORM
                                                        7.   1.1,1-TRICHLOROETHANE
                                                        8.   1,2-DICHLORORETHYLENE
                                                        9.   CARBON TETRACHLORIDE
                                                        10.  BENZENE
                                                        11.  FLOUROBENZENE (INT. STD.)
                                                        12.  TRICHLOROETHYENE
                                                        13.1,2-DICHLOROPROPAKE
                                                        14.  BROMODICHLOROHETHANE
                                                        IS.  TOLUENE
                                                        16.  BROHOCHLOROPROPANE INT. STD.)
                                                        17.  DIBROMOCHLOROHETHANE
                                                        18.  TETRACHLOROETHYLENE
                                                        19.  CHLOR08ENZENE
                                                        20.  ETHYLBENZENE
                                                        21.  1,3-XYLENE
                                                        22.  BROMOFORM
                                                        23.  BROMOBENZENE
                                                        24.  1.4-DICHLOROBENZENE
                                                        25.  1,2,4-TRICHLOROBENZENE
                                                        26.  NAPHTHALENE
                                                                8360A  -  52
                                                                                       Revision 1
                                                                                  September 1994

-------
                   FIGURE 9.
               LOW SOILS  IMPINGER
                                —'
 PURGE INLET FITTING
 SAMPLE OUTLET PITTING
3- • 6mm 0 D CLASS TUBING
                                     SEPTUM
                                        CAP
             40mi VIAL
                 8260A - 53
    Revision 1
September 1994

-------
                                       METHOD  8260A
VOLATILE ORGANIC COMPOUNDS BY GAS CHROMATOGRAPHY/MASS  SPECTROMETRY  (GC/MS)
                              CAPILLARY  COLUMN TECHNIQUE
                    Purge-and-trap
                               7.1
                              Select
                             procedure
                           for introducing
                            sample into
                              GC/MS.
 Direct
Injection
    ^
                           7.2 Set GC/MS
                              operating
                             conditions.
                             7.3.1 Tune
                           GC/MS system
                              with BFB.
                           7.3.2 Assemble
                            purge-and-trap
                          device and prepare
                         calibration standards
                           7.3.2.1 Perform
                            purge-and-trap
                              analysis.
                                        8260A  -  54
7.3.4 Calculate
    RFs for
   5 SPCCs.
               7.3.5 Calculate
                %RSD of RF
                 for CCCs.
               7.4 Perform
                calibration
               verification.
            7.5 Perform GC/MS
              analysis utilizing
               Methods 5030
                 or 8260.
               7.6.1  Identify
                analytes by
              comparing the
            sample and standard
               mass spectra.
                                                         7.6.2 Calculate the
                                                          concentration of
                                                           each identified
                                                              analyte.
                                                           7.6.2.3 Report
                                                             all results.
                                                           f    Stop    J
                                      Revision  1
                                 September  1994

-------
00
*o
-4
o

-------
                                 METHOD 8270A

                       SEMIVOLATILE ORGANIC  COMPOUNDS  BY
   GAS CHROMATOGRAPHY/MASS SPECTROMETRY (GC/MS); CAPILLARY COLUMN TECHNIQUE


1.0   SCOPE AND APPLICATION

      1.1   Method 8270 is used to determine the concentration of semivolatile
organic compounds in extracts prepared from all types of solid waste matrices,
soils, and ground water.  Direct  injection  of  a sample may be used in limited
applications.  The following compounds can be determined by this method:
Compounds
        Appropriate Preparation Techniques

CAS No8   3510     3520  3540  3550  3580
Acenaphthene
Acenaphthene-d10 (I.S.)
Acenaphthylene
Acetophenone
2-Acetyl ami nof 1 uorene
l-Acetyl-2-thiourea
Aldrin
2-Aminoanthraquinone
Aminoazobenzene
4-Aminobiphenyl
Anilazine
Aniline
o-Anisidine
Anthracene
Aramite
Aroclor - 1016
Aroclor - 1221
Aroclor - 1232
Aroclor - 1242
Aroclor - 1248
Aroclor - 1254
Aroclor - 1260
Azinphos-methyl
Barban
Benzidine
Benzoic acid
Benz(a)anthracene
Benzo (b)fl uoranthene
Benzo ( k) fl uoranthene
Benzo (g,h,i)perylene
Benzo(a)pyrene
p-Benzoquinone
Benzyl alcohol
83-32-9

208-96-8
98-86-2
53-96-3
591-08-2
309-00-2
117-79-3
60-09-3
92-67-1
101-05-3
62-53-3
90-04-0
120-12-7
140-57-8
12674-11-2
11104-28-2
11141-16-5
53469-21-9
12672-29-6
11097-69-1
11096-82-5
86-50-0
101-27-9
92-87-5
65-85-0
56-55-3
205-99-2
207-08-9
191-24-2
50-32-8
106-51-4
100-51-6
X
X
X
X
X
LR
X
X
X
X
X
X
X
X
HS(43)
X
X
X
X
X
X
X
HS(62)
LR
CP
X
X
X
X
X
X
OE
X
X
X
X
NO
ND
ND
X
ND
ND
ND
ND
X
ND
X
ND
X
X
X
X
X
X
X
ND
ND
CP
X
X
X
X
X
X
ND
X
X
X
X
ND
ND
ND
X
ND
ND
ND
ND
ND
ND
X
ND
X
X
X
X
X
X
X
ND
ND
CP
ND
X
X
X
X
X
ND
ND
X
X
X
ND
ND
ND
X
ND
ND
ND
ND
X
ND
X
ND
X
X
X
X
X
X
X
ND
ND
CP
X
X
X
X
X
X
ND
X
X
X
X
X
X
LR
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
LR
CP
X
X
X
X
X
X
X
X
                                   8270A -  1
                                Revision 1
                                 July 1992

-------
Appropriate Preparation Techniaues
Compounds
a-BHC
/3-BHC
5-BHC
y-BHC (Lindane)
Bis(2-chloroethoxy)methane
Bis(2-chloroethyl) ether
Bis(2-chloroisopropyl) ether
Bis(2-ethylhexyl) phthalate
4-Bromophenyl phenyl ether
Bromoxynil
Butyl benzyl phthalate
Captafol
Captan
Carbaryl
Carbofuran
Carbophenothion
Chlordane
Chlorfenvinphos
4-Chloroaniline
Chi orobenzi late
5-Chl oro-2-methyl ani 1 i ne
4-Chloro-3-methyl phenol
3-(Chloromethyl )pyridine
hydrochloride
1-Chloronaphthalene
2-Chloronaphthalene
2-Chlorophenol
4-Chlorophenyl phenyl ether
Chrysene
Chrysene-d12 (I.S.)
Coumaphos
p-Cresidine
Crotoxyphos
2-Cyclohexyl-4,6-dinitrophenol
4,4'-DDD
4,4'-DDE
4, 4 '-DDT
Oemeton-0
Demeton-S
Diallate (cis or trans)
2,4-Diaminotoluene
Dibenz(a,j)acridine
Dtbenz(a,h) anthracene
Oibenzofuran
Dibenzo(a,e)pyrene
Di-n-butyl phthalate

CAS Noa
319-84-6
319-85-7
319-86-8
58-89-9
111-91-1
111-44-4
108-60-1
117-81-7
101-55-3
1689-84-5
85-68-7
2425-06-1
133-06-2
63-25-2
1563-66-2
786-19-6
57-74-9
470-90-6
106-47-8
510-15-6
95-79-4
59-50-7

6959-48-4
90-13-1
91-58-7
95-57-8
7005-72-3
218-01-9

56-72-4
120-71-8
7700-17-6
131-89-5
72-54-8
72-55-9
50-29-3
298-03-3
126-75-0
2303-16-4
95-80-7
224-42-0
53-70-3
132-64-9
192-65-4
84-74-2
8270A - 2
3510
X
X
X
X
X
X
X
X
X
X
X
HS(55)
HS(40)
X
X
X
X
X
X
X
X
X

X
X
X
X
X
X
X
X
X
X
X
X
X
X
HS(68)
X
X
DC,OE(42)
X
X
X
ND
X

3520
X
X
X
X
X
X
X
X
X
ND
X
ND
ND
ND
ND
ND
X
ND
ND
ND
NO
X

ND
X
X
X
X
X
X
ND
ND
ND
ND
X
X
X
ND
ND
ND
ND
ND
X
X
ND
X

3540
X
X
X
X
X
X
X
X
X
ND
X
ND
ND
ND
ND
ND
X
ND
ND
ND
ND
X

ND
X
X
X
X
X
X
ND
ND
ND
ND
X
X
X
ND
ND
ND
ND
ND
X
ND
ND
X

3550
X
X
X
X
X
X
X
X
X
ND
X
ND
ND
ND
ND
ND
X
ND
ND
ND
NO
X

ND
X
X
X
X
X
X
ND
ND
ND
ND
X
X
X
ND
ND
ND
ND
ND
X
X
ND
X
Revi
3580
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X

X
X
X
X
X
X
X
X
X
X
LR
X
X
X
X
X
X
X
X
X
X
X
X
sion 1
July 1992

-------
Compounds
Dichlone
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
l,4-Dichlorobenzene-d4 (I.S)
3,3'-Dichlorobenzidine
2,4-Dichlorophenol
2,6-Dichlorophenol
Dichlorovos
Dicrotophos
Oieldrin
Diethyl phthalate
Diethylstilbestrol
Diethyl sulfate
Dimethoate
3,3'-Dimethoxybenzidine
Dimethyl ami noazobenzene
7,12-Dimethylbenz(a)-
anthracene
3,3' -Dimethylbenzidine
a, a -Dimethyl phenethyl ami ne
2,4-Dimethylphenol
Dimethyl phthalate
1,2-Dinitrobenzene
1,3-Dinitrobenzene
1,4-Dinitrobenzene
4,6-Dinitro-2-methylphenol
2,4-Dinitrophenol
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Dinocap
Dinoseb
Diphenylamine
5,5-Diphenylhydantoin
1,2-Diphenylhydrazine
Di-n-octyl phthalate
Disulfoton
Endosulfan I
Endosulfan II
Endosulfan sulfate
Endrin
Endrin aldehyde
Endrin ketone
EPN
Ethion
Ethyl carbamate


t
CAS Noa
117-80-6
95-50-1
541-73-1
106-46-7

91-94-1
120-83-2
87-65-0
62-73-7
141-66-2
60-57-1
84-66-2
56-53-1
64-67-5
60-51-5
119-90-4
60-11-7

57-97-6
119-93-7
122-09-8
105-67-9
131-11-3
528-29-0
99-65-0
100-25-4
534-52-1
51-28-5
121-14-2
606-20-2
39300-45-3
88-85-7
122-39-4
57-41-0
122-66-7
117-84-0
298-04-4
959-98-8
33213-65-9
1031-07-8
72-20-8
7421-93-4
53494-70-5
2104-64-5
563-12-2
51-79-6
8270A - 3

yopropriate Preoaration Techniaues
3510
OE
X
X
X
X
X
X
X
X
X
X
X
AW,OS(67)
LR
HE,HS(31)
X
X

CP(45)
X
ND
X
X
X
X
HE(14)
X
X
X
X
CP,HS(28)
X
X
X
X
X
X
X
X
X
X
X
X
X
X
DC(28)
3520
ND
X
X
X
X
X
X
ND
ND
ND
X
X
ND
ND
ND
ND
ND

ND
ND
ND
X
X
ND
ND
ND
X
X
X
X
ND
ND
X
ND
X
X
ND
X
X
X
X
X
X
ND
ND
ND
3540
ND
X
X
X
X
X
X
ND
ND
ND
X
X
ND
ND
ND
ND
ND

ND
ND
ND
X
X
ND
ND
ND
X
X
X
X
ND
ND
X
ND
X
X
ND
X
X
X
X
X
ND
ND
ND
ND
3550 3580
ND
X
X
X
X
X
X
ND
ND
ND
X
X
ND
ND
ND
ND
ND

ND
ND
ND
X
X
ND
ND
ND
X
X
X
X
ND
ND
X
ND
X
X
ND
X
X
X
X
X
X
ND
ND
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
LR
X
LR
X

CP
X
X
X
X
X
X
X
X
X
X
X
CP
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
Revision 1



July
1992

-------
Compounds
Ethyl methanesulfonate
Famphur
Fensulfothion
Fenthion
Fluchloralin
Fluoranthene
Fluorene
2-Fluorobiphenyl (surr.)
2-Fluorophenol (surr.)
Heptachlor
Heptachlor epoxide
Hexachlorobenzene
Hexachlorobutadiene
Hexachl orocycl opentadi ene
Hexachloroethane
Hexachl orophene
Hexachl oropropene
Hexamethyl phosphorami de
Hydroquinone
Indeno(l,2,3-cd)pyrene
Isodrin
Isophorone
Isosafrole
Kepone
Leptophos
Malathion
Maleic anhydride
Mestranol
Methapyrilene
Methoxychlor
3-Methylcholanthrene
4,4'-Methylenebis(2-chloranil
Methyl methanesulfonate
2-Methylnaphthalene
Methyl parathion
2-Methyl phenol
3-Methyl phenol
4-Methyl phenol
Mevinphos
Mexacarbate
Mirex
Monocrotophos
Naled
Naphthalene
Naphthalene-dg (I.S.)
1,4-Naphthoquinone
1-Naphthylamine

i
CAS Noa
62-50-0
52-85-7
115-90-2
55-38-9
33245-39-5
206-44-0
86-73-7
321-60-8
367-12-4
76-44-8
1024-57-3
118-74-1
87-68-3
77-47-4
67-72-1
70-30-4
1888-71-7
680-31-9
123-31-9
193-39-5
465-73-6
78-59-1
120-58-1
143-50-0
21609-90-5
121-75-5
108-31-6
72-33-3
91-80-5
72-43-5
56-49-5
ine) 101-14-4
66-27-3
91-57-6
298-00-0
95-48-7
108-39-4
106-44-5
7786-34-7
315-18-4
2385-85-5
6923-22-4
300-76-5
91-20-3

130-15-4
134-32-7
8270A - 4
\DDrooriate Preoaration Terhnioues
3510
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
AW,CP(62)
X
X
NO
X
X
X
DC(46)
X
X
HS(5)
HE
X
X
X
X
OE,OS(0)
X
X
X
X
X
X
X
HE,HS(68)
X
HE
X
X
X
X
OS(44)

3520
ND
NO
ND
NO
ND
X
X
X
X
X
X
X
X
X
X
ND
ND
ND
ND
X
ND
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
X
X
ND
ND

3540
ND
ND
ND
ND
ND
X
X
X
X
X
X
X
X
X
X
ND
ND
ND
ND
X
ND
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
X
X
ND
ND

3550
ND
ND
ND
ND
ND
X
X
X
X
X
X
X
X
X
X
ND
ND
ND
ND
X
ND
X
ND
ND
ND
ND
ND
NO
ND
ND
ND
ND
ND
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
X
X
ND
ND
Revi
3580
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
CP
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
LR
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
si on 1
July 1992

-------
Appropriate Preoaration Techniaue<;
Compounds
2-Naphthylamine
Nicotine
5-Nitroacenaphthene
2-Nitroaniline
3-Nitroaniline
4-Nitroaniline
5-Nitro-o-anisidine
Nitrobenzene
Nitrobenzene-dr (surr.)
4-Nitrobiphenyl
Nitrofen
2-Nitrophenol
4-Nitrophenol
Nitroquinoline-1 -oxide
N-Nitrosodibutylamine
N-Nitrosodi ethyl ami ne
N-Nitrosodimethylamine
N-Nitrosodiphenylamine
N-Nitrosodi-n-propylamine
N-Ni trosomethyl ethyl ami ne
N-Nitrosomorphol ine
N-Nitrosopipendine
N-Nitrosopyrrol idine
5-Nitro-o-toluidine
Octamethyl pyrophosphoramide
4,4'-Oxydiani1ine
Parathion
Pentachlorobenzene
Pentachloronitrobenzene
Pentachlorophenol
Pery1ene-d12 (I.S.)
Phenacetin
Phenanthrene
Phenanthrene-d10 (I.S.)
Phenobarbital
Phenol
Phenol -d6 (surr.)
1,4-Phenylenediamine
Phorate
Phosalone
Phosmet
Phosphamidon
Phthalic anhydride
2-Picoline
Piperonyl sulfoxide
Pronamide
Propylthiouracil

CAS No8
91-59-8
54-11-5
602-87-9
88-74-4
99-09-2
100-01-6
99-59-2
98-95-3
92-93-3
1836-75-5
88-75-5
100-02-7
56-57-5
924-16-3
55-18-5
62-75-9
86-30-6
621-64-7
10595-95-6
59-89-2
100-75-4
930-55-2
99-55-8
152-16-9
101-80-4
56-38-2
608-93-5
82-68-8
87-86-5

62-44-2
85-01-8

50-06-6
108-95-2

106-50-3
298-02-2
2310-17-0
732-11-6
13171-21-6
85-44-9
109-06-8
120-62-7
23950-58-5
51-52-5
8270A - 5
3510
X
DE(67)
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
ND
X
X
X
LR
X
X
X
X
X
X
X
X
X
X
DC(28)
DC(28)
X
X
HS(65)
HS(15)
HE(63)
CP,ME(1)
ND
X
X
LR

3520
ND
ND
ND
X
X
X
ND
X
X
ND
ND
X
X
ND
ND
ND
X
X
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
X
X
ND
X
X
ND
X
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND

3540
ND
ND
ND
ND
ND
ND
ND
X
X
ND
ND
X
X
ND
ND
ND
X
X
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
X
X
ND
X
X
ND
X
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND

3550
ND
ND
ND
X
X
X
ND
X
X
ND
ND
X
X
ND
ND
ND
X
X
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
X
X
ND
X
X
ND
X
X
ND
ND
ND
NO
ND
ND
ND
ND
ND
ND
Revi
3580
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
LR
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
CP
ND
X
X
LR
sion 1
July 1992

-------
Compounds
        Appropriate Preparation Techniques

CAS Noa   3510     3520  3540  3550  3580
Pyrene
Pyridine
Resorcinol
Safrole
Strychnine
Sul fall ate
Terbufos
Terphenyl-du(surr.)
1,2,4,5-Tetrachlorobenzene
2,3,4,6-Tetrachlorophenol
Tetrachlorvinphos (Stirophos)
Tetraethyl pyrophosphate
Thionazine
Thiophenol (Benzenethiol )
Toluene diisocyanate
o-Toluidine
Toxaphene
2,4,6-Tribromophenol (surr.)
1,2,4-Trichlorobenzene
2,4,5-Trichlorophenol
2,4,6-Trichlorophenol
Trifluralin
2,4,5-Trime.thylani"! ine
Trimethyl phosphate
1,3,5-Trinitrobenzene
Tris(2,3-dibromopropyl) phosphate
Tri-p-tolyl phosphate
0,0,0-Triethyl phosphorothioate
129-00-0
110-86-1
108-46-3
94-59-7
60-41-3
95-06-7
13071-79-9

95-94-3
58-90-2
961-11-5
107-49-3
297-97-2
108-98-5
584-84-9
95-53-4
8001-35-2

120-82-1
95-95-4
88-06-2
1582-09-8
137-17-7
512-56-1
99-35-4
126-72-7
78-32-0
126-68-1
X
ND
DC.OE(IO)
X
AW,OS(55)
X
X
X
X
X
X
X
X
X
HE(6)
X
X
X
X
X
X
X
X
HE(60)
X
X
X
X
X
ND
NO
ND
ND
ND
ND
X
ND
ND
ND
ND
ND
ND
ND
ND
X
X
X
X
X
ND
ND
ND
ND
ND
ND
ND
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
X
X
X
ND
X
ND
ND
ND
ND
ND
ND
ND
X
ND
ND
ND
ND
ND
ND
X
ND
ND
ND
ND
ND
ND
ND
ND
X
X
X
X
X
ND
ND
ND
ND
ND
ND
ND
X
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
x
X
X
X
LR
X
X
a    Chemical Abstract Service Registry Number.

AW = Adsorption to walls of glassware during extraction and storage.
CP = Nonreproducible chromatographic performance.
DC = Unfavorable  distribution  coefficient  (number  in  parenthesis  is  percent
     recovery).
HE = Hydrolysis during  extraction accelerated  by  acidic or  basic  conditions
     (number in parenthesis is percent recovery).
HS = Hydrolysis during storage (number in parenthesis is percent stability).
LR = Low response.
ND = Not determined.
OE = Oxidation  during  extraction accelerated  by basic conditions  (number  in
     parenthesis  is percent recovery).
OS = Oxidation during storage (number in parenthesis is percent stability).
X  = Greater than 70 percent recovery by this technique.

Percent Stability =  Average Recovery (Day 7)  x 100/Average  Recovery (Day  0).
                                   8270A -  6
                                Revision 1
                                 July 1992

-------
      1.2   Method 8270 can  be used to quantitate most neutral,  acidic, and basic
organic compounds that are  soluble  in  methylene chloride and capable of being
eluted without derivatization as sharp peaks from a gas chromatographic fused-
silica capillary column coated with a slightly polar silicone.  Such compounds
include  polynuclear   aromatic   hydrocarbons,   chlorinated  hydrocarbons  and
pesticides, phthalate esters, organophosphate esters, nitrosamines, haloethers,
aldehydes,   ethers,  ketones, anilines,  pyridines,  quinolines,  aromatic  nitro
compounds,   and  phenols,  including nitrophenols.   See Table  1  for a  list  of
compounds and their characteristic ions that have been evaluated on the specified
GC/MS system.

      1.3   The  following  compounds  may  require  special  treatment  when  being
determined by this method.   Benzidine  can be subject to oxidative losses during
solvent  concentration.   Also,  chromatography  is  poor.    Under the  alkaline
conditions of the extraction step, a-BHC, yBHC, Endosulfan I and II, and Endrin
are subject to decomposition.  Neutral extraction should  be performed if these
compounds  are  expected.    Hexachlorocyclopentadiene  is  subject   to  thermal
decomposition in the inlet of the  gas chromatograph,  chemical reaction in acetone
solution, and photochemical  decomposition.  N-nitrosodimethylamine is difficult
to separate from the solvent under the chromatographic conditions described.  N-
nitrosodiphenylamine decomposes in the gas chromatographic inlet and cannot  be
separated   from   diphenylamine.      Pentachlorophenol,   2,4-dinitrophenol,
4-nitrophenol, 4,6-dinitro-2-methylphenol,4-chloro-3-methylphenol, benzoicacid,
2-nitroaniline,  3-nitroaniline, 4-chloroaniline,  and benzyl  alcohol are subject
to erratic  chromatographic  behavior, especially if the GC  system  is contaminated
with high boiling material.

      1.4   The estimated  quantitation limit (EQL) of Method  8270  for determining
an individual compound is approximately 1 mg/kg (wet weight) for soil/sediment
samples, 1-200 mg/kg  for wastes (dependent on matrix and method of preparation),
and  10  M9/L   for  ground  water  samples  (see  Table  2).    EQts  will  be
proportionately  higher for  sample  extracts  that  require  dilution to  avoid
saturation of the detector.

      1.5   This  method  is  restricted  to  use  by or  under  the  supervision  of
analysts experienced  in  the use  of gas  chromatograph/mass  spectrometers and
skilled in  the interpretation of mass spectra.  Each analyst  must  demonstrate the
ability to generate acceptable results with this method.


2.0   SUMMARY OF METHOD

      2.1   Prior  to  using  this  method,  the  samples  should  be  prepared for
chromatography  using  the  appropriate sample preparation and  cleanup methods.
This  method describes  chromatographic  conditions that  will  allow  for  the
separation  of the  compounds   in  the extract  and  for  their qualitative  and
quantitative analysis by mass spectrometry.


3.0   INTERFERENCES

      3.1   Raw GC/MS data from  all blanks, samples,  and  spikes  must be evaluated
for interferences. Determine if the  source of interference is in  the preparation


                                  8270A - 7                         Revision 1
                                                                     July 1992

-------
and/or cleanup  of the  samples and  take corrective  action  to  eliminate  the
problem.

     3.2   Contamination by carryover can occur whenever high-concentration and
low-concentration samples are sequentially analyzed.  To reduce carryover,  the
sample syringe must  be  rinsed  out between samples with solvent.   Whenever an
unusually concentrated  sample  is  encountered,  it  should  be followed  by  the
analysis of solvent to check for cross contamination.


4.0  APPARATUS AND MATERIALS

     4.1   Gas chromatograph/mass spectrometer system

           4.1.1  Gas chromatograph  - An analytical  system complete with  a
     temperature-programmable gas chromatograph suitable for splitless injection
     and  all  required accessories, including syringes, analytical columns,  and
     gases.   The  capillary  column  should be directly  coupled to  the source.

           4.1.2  Column - 30 m x 0.25 mm  ID (or 0.32  mm ID) 1 jum film thickness
     silicone-coated  fused-silica capillary  column  (J&W  Scientific DB-5  or
     equivalent).

           4.1.3  Mass spectrometer  -  Capable of scanning from  35  to 500  amu
     every  1 sec  or less, using  70 volts (nominal)  electron energy  in  the
     electron impact ionization mode.  The mass spectrometer  must be capable of
     producing  a  mass  spectrum for decafluorotriphenylphosphine (DFTPP) which
     meets all of the criteria  in Table 3 when  1 ML of the GC/MS tuning standard
     is  injected  through  the GC  (50  ng of DFTPP).

           4.1.4  GC/MS interface - Any GC-to-MS  interface  that gives acceptable
     calibration  points at 50 ng  per  injection for each compound  of interest and
     achieves acceptable  tuning  performance criteria  may be  used.

           4.1,5  Data system - A computer system must be interfaced to the mass
     spectrometer.  The system  must allow the continuous acquisition and storage
     on  machine-readable media  of all  mass  spectra  obtained  throughout,  the
     duration of  the chromatographic program.   The computer must have software
     that can search any  GC/MS data  file for  ions of a specific mass and that
     can  plot such ion abundances versus  time or scan number.  This type of plot
     is defined as an Extracted Ion Current Profile (EICP).   Software must also
     be  available that allows  integrating the abundances  in any EICP between
     specified  time or scan-number  limits.   The  most recent  version  of  the
     EPA/NIST Mass  Spectral  Library  should also  be available.

     4.2   Syringe - 10 nl.

     4.3   Volumetric  flasks,  Class A - Appropriate  sizes  with  ground glass
stoppers.

     4.4   Balance - Analytical, 0.0001  g.

     4.5   Bottles - glass with Teflon-lined screw caps or crimp tops.


                                   8270A  - 8                        Revision  1
                                                                     July  1992

-------
5.0  REAGENTS

     5.1   Reagent grade chemicals  shall be used in all  tests.  Unless otherwise
indicated,  it is intended that  all  reagents shall conform to the specifications
of the  Committee on Analytical  Reagents  of the American Chemical Society, where
such specifications are  available.   Other  grades may  be used,  provided it is
first ascertained that the reagent  is  of sufficiently high purity to permit its
use without lessening the accuracy of the determination.

     5.2   Organic-free reagent water - All  references to water in this method
refer to organic-free reagent water, as defined in Chapter One.

     5.3   Stock  standard  solutions (1000  mg/L)  -  Standard solutions  can be
prepared from pure standard materials or purchased as certified solutions.

           5.3.1  Prepare stock standard solutions by accurately weighing about
     0.0100  g  of pure material.   Dissolve  the material in  pesticide quality
     acetone  or  other  suitable  solvent  and  dilute  to volume  in  a  10  ml
     volumetric  flask.   Larger volumes can be used at the  convenience of the
     analyst.  When compound purity is assayed  to be 96% or greater, the weight
     may be used without correction to calculate the concentration of the stock
     standard.    Commercially  prepared stock  standards may be  used  at  any
     concentration  if  they  are  certified   by  the   manufacturer  or   by  an
     independent  source.

           5.3.2  Transfer the  stock standard solutions into  bottles with Teflon
     lined  screw-caps.   Store  at 4°C and protect from light.   Stock standard
     solutions  should  be  checked  frequently  for  signs   of  degradation  or
     evaporation, especially just prior to preparing calibration standards from
     them.

           5.3.3  Stock  standard  solutions  must be  replaced after 1  year or
     sooner  if  comparison with  quality control  check  samples  indicates  a
     problem.

     5.4   Internal standard solutions - The  internal standards recommended are
1,4-dichlorobenzene-d,,    naphthalene-d8,  acenaphthene-d10,    phenanthrene-d^,
chrysene-d12, and perylene-d12.  Other compounds may be used as  internal standards
as long as  the  requirements given in Section 7.3.2 are met.   Dissolve 0.200 g of
each compound  with  a  small volume  of  carbon  disulfide.   Transfer to  a 50 ml
volumetric flask and dilute to  volume  with methylene chloride so that the final
solvent is approximately 20% carbon disulfide.   Most of the compounds are also
soluble  in  small  volumes  of  methanol,  acetone,  or  toluene,  except  for
perylene-d<2.     The   resulting   solution  will  contain  each  standard  at  a
concentration  of  4,000 ng//j,L.    Each  1 mL  sample extract  undergoing analysis
should be spiked with 10 /LtL (5 \il if a  2 \il injection  is used) of the internal
standard solution,  resulting in  a  concentration of 40 ng//iL of each internal
standard.  Store at 4°C or less when not being  used.

     5.5   -GC/MS  tuning  standard -  A methylene chloride  solution containing
50 ng/^L (25 ng/jiL if a 2 \ii injection  is used) of decafluorotriphenylphosphine
(DFTPP) should be prepared.  The standard should also contain 50 ng//nL each of
4,4'-DDT, pentachlorophenol, and benz.idine to  verify injection port inertness and
GC column perfo'rmance.  Store at 4°C or less  when  not being  used.

                                   8270A -  9                         Revision 1
                                                                     July  1992

-------
     5.6   Cal ibration standards - A minimum of five cal ibration standards should
be prepared.  One of the calibration standards should be at  a concentration near,
but above,  the method detection limit; the others should correspond to the range
of concentrations found in  real samples  but should  not  exceed the working range
of the  GC/MS  system.  Each standard should contain each  analyte for detection by
this method  (e.g.  some  or all  of  the compounds  listed  in Table  1  may  be
included).    Each  1  ml aliquot of  calibration  standard should be  spiked  with
10 pi (5 [iL if a  2 jiL injection is used)  of the internal standard solution prior
to analysis.  All  standards  should be  stored at  -10°C to  -20°C  and  should  be
freshly prepared once a year,  or  sooner if check standards indicate a problem.
The daily calibration standard should be prepared weekly  and stored at  4°C.

     5.7   Surrogate  standards  -  The   recommended  surrogate  standards  are
phenol-d6,    2-fluorophenol,     2,4,6-tribromophenol,     nitrobenzene-ds,
2-fluorobiphenyl, and p-terphenyl-d.4.   See Method 3500 for the instructions on
preparing the surrogate standards.  Determine what concentration should be in the
blank extracts  after all extraction,  cleanup,  and concentration steps.   Inject
this concentration into the GC/MS to  determine recovery of surrogate standards
in all  blanks,  spikes,  and  sample extracts.  Take into  account all dilutions of
sample extracts.

     5.8   Matrix spike standards - See Method 3500 for instructions on preparing
the matrix spike standard.  Determine what concentration should be in the blank
extracts after all  extraction, cleanup,  and concentration  steps.   Inject this
concentration into the  GC/MS to determine recovery of surrogate standards in all
matrix spikes.   Take into account all dilutions of sample extracts.

     5.9   Acetone, hexane, methylene chloride, isooctane,  carbon disulfide,
toluene, and other appropriate solvents - Pesticide quality or equivalent


6.0  SAMPLE  COLLECTION,  PRESERVATION, AND HANDLING

     6.1   See  the  introductory  material to  this chapter,  Organic  Analytes,
Section 4.1.
7.0   PROCEDURE

      7.1   Sample preparation - Samples must be prepared by one of the following
methods prior  to GC/MS analysis.

      Matrix             Methods
      Water             3510, 3520
      Soil/sediment      3540, 3550
      Waste             3540, 3550, 3580

           7.1.1  Direct  injection  -  In  very   limited   applications  direct
      injection of the sample  into  the GC/MS system with a  10>L syringe may be
      appropriate.  The detection limit is very high (approximately 10,000 M9/L);
      therefore, it   is  only  permitted  where  concentrations  in  excess  of
      10,000  /zg/L are expected.   The  system must  be  calibrated by  direct
      injection.


                                   8270A - 10                        Revision  1
                                                                     July 1992

-------
     7.2   Extract cleanup - Extracts may be cleaned  up  by any of the following
methods prior to GC/MS analysis.
     Compounds
     Phenols
     Phthalate esters
     Nitrosamines
     Organochlorine pesticides & PCBs
     Nitroaromatics and cyclic ketones
     Polynuclear aromatic hydrocarbons
     Haloethers
     Chlorinated hydrocarbons
     Organophosphorus pesticides
     Petroleum waste
     All priority pollutant base,
         neutral, and acids
            Methods
            3630, 3640, 8040a
            3610, 3620, 3640
            3610, 3620, 3640
            3620, 3660
            3620, 3640
            3611, 3630, 3640
            3620, 3640
            3620, 3640
            3620
            3611, 3650

            3640
"Method 8040  includes a derivatization technique followed  by GC/ECD analysis,
interferences are encountered on GC/FID.

     7.3   Initial calibration - The recommended GC/MS operating conditions:
                                        if
     Mass range:
     Scan time:
     Initial temperature:
     Temperature program:
     Final temperature:

     injector temperature:
     Transfer line temperature:
     Source temperature:
     Injector:
     Sample volume:
     Carrier gas:
35-500 amu
1 sec/scan
40°C,  hold for 4 minutes
40-270°C at 10°C/min
270°C,  hold until benzo[g,h,i]perylene has
eluted
250-300°C
250-300°C
According to manufacturer's specifications
Grob-type, splitless
1-2 ML
Hydrogen  at  50 cm/sec  or  helium  at  30
cm/sec
           7.j.l  Each GC/MS system must  be hardware-tuned to meet the criteria
      in Table 3 for a 50 ng  injection of DFTPP.   Analyses  should not begin until
      all   these  criteria  are   met.     Background   subtraction   should   be
      straightforward  and designed only to eliminate column bleed or instrument
      background  ions.  The  GC/MS tuning  standard should also be used to ass.ess
      GC column performance  and injection  port inertness.  Degradation of DOT to
      DDE and ODD should not exceed 20%.   Benzidine and pentachlorophenol should
      be  present at  their  normal  responses, and  no  peak  tailing  should  be
      visible.  If degradation is excessive and/or poor chromatography is noted,
      the injection port may require cleaning.  It may also be necessary to break
      off the first 6-12 in. of  the  capillary column.

           7.3.2  The internal standards selected in Section 5.1 should permit
      most  of the components  of  interest in  a  chromatogram to  have retention
      times of 0.80-1.20 relative to one of the internal  standards.  Use the base
      peak  ion   from  the  specific  internal   standard  as the  primary  ion  for
      quantitation  (see Table  1).  If interferences  are  noted, use the next most
                                  8270A - 11
                                Revision 1
                                 July 1992

-------
intense ion as  the  quantitation  ion  (i.e.  for l,4-dichlorobenzene-d4 use
m/z 152 for quantitation).

      7.3.3  Analyze 1 pi of each  calibration standard (containing internal
standards) and tabulate the area  of the primary characteristic ion against
concentration  for each compound (as  indicated in Table 1).  Figure 1 shows
a chromatogram of a calibration standard containing base/neutral  and acid
analytes.   Calculate response factors (RFs) for each compound as  follows:

                         RF =  (AxCit)/(A,.Cx)

where:

      Ax     =     Area of the characteristic ion  for the  compound being
                   measured.

      A-s    =     Area of the characteristic ion for the specific  internal
                   standard.
      Cjs     =      Concentration of the specific internal standard

      Cx      =      Concentration  of the compound  being  measured (ng//nL).

      7.3.4   The average RF should be calculated for each  compound.   The
percent relative standard  deviation  (%RSD =  100[SD/RF])  should  also be
calculated for each compound.  The %RSD should be less  than  30% for  each
compound.  However, the %RSO for each  individual Calibration Check Compound
(CCC) (see Table 4) must be  less  than  30%.   The  relative retention  times
of each compound in each calibration  run should agree within 0.06 relative
retention time  units.   Late-eluting  compounds  usually have much  better
agreement.

      7.3.5   A  system performance check must be  performed  to ensure  that
minimum average  RFs  are met before  the calibration  curve is  used.   For
semivolatiles,   the  System Performance  Check Compounds  (SPCCs) are:  N-
nitroso-di-n-propylamine;  hexachlorocyclopentadiene;  2,4-dinitro-phenol ;
and 4-nitrophenol .   The minimum acceptable average  RF for these compounds
SPCCs is 0.050.   These SPCCs typically have very low RFs (0.1-0.2) and tend
to decrease  in response as the chromatographic system begins to deteriorate
or the standard  material begins to deteriorate.  They are usually the first
to  show  poor   performance.    Therefore,  they  must  meet  the  minimum
requirement when the system  is calibrated.

7.4   Daily  GC/MS calibration

      7.4.1   Prior  to analysis of samples,  the  GC/MS  tuning standard  must
be analyzed. A  50 ng injection of DFTPP must  result in a mass spectrum for
DFTPP which meets the criteria given  in Table 3.  These criteria must be
demonstrated during each 12  hour shift.

      7.4.2   A  calibration standard(s)  at mid-concentration,  containing
each  compound  of  interest,  including all  required  surrogates, must  be
performed every 12  hours during analysis.  Compare the response factor data
from the standards  every 12 hours with the  average response factor from the

                             8270A -  12                        Revision 1
                                                                July  1992

-------
initial calibration  for a specific  instrument  as per the  SPCC  (Section
7.4.3) and CCC (Section 7.4.4) criteria.

      7.4.3  System  Performance   Check  Compounds  (SPCCs):    A  system
performance check must  be  made during every 12 hour shift.   If  the  SPCC
criteria  are  met,  a comparison  of  response  factors  is  made  for  all
compounds.   This is the  same  check  that  is  applied during  the  initial
calibration.  If the minimum  response factors are not met, the system must
be evaluated, and corrective action  must  be  taken  before sample  analysis
begins.   The  minimum RF for  semivolatile SPCCs is 0.050.   Some  possible
problems   are  standard  mixture   degradation,   injection   port   inlet
contamination, contamination  at the front end of the analytical column, and
active sites  in the column or chromatographic system.  This check must be
met before analysis begins.

      7.4.4  Calibration  Check   Compounds  (CCCs):     After  the  system
performance check  is met,  CCCs  listed  in  Table 4  are used to check the
validity  of  the  initial calibration.   Calculate  the  percent difference
using:

                                  RF,   - RFC
                  % Difference = 	 x 100
                                     RF,
where:
      RFj     =     Average  response  factor  from   initial   calibration
                   (Section  7.3).
      RFC     =     Response factor from current verification check standard.

      If the percent difference for any compound is  greater  than  20,  the
laboratory should consider this a warning limit.  If the percent difference
for each CCC  is  less  than 30%,  the initial calibration is assumed  to  be
valid.  If  the criterion  is  not  met (> 30% difference) for  any one  CCC,
corrective action must be taken.   Problems similar to  those  listed  under
SPCCs could  affect  this  criterion.   If no  source  of the problem can  be
determined  after corrective  action  has  been  taken,  a  new  five-point
calibration must be generated.  This  criterion must  be met before  sample
analysis begins.

      7.4.5   The  internal  standard  responses  and  retention  times in  the
calibration check standard must be  evaluated  immediately  after or during
data acquisition.  If  the  retention  time for any internal  standard  changes
by more than 30 seconds from the las-t daily calibration (Section 7.4),  the
chromatographic system must be inspected for  malfunctions  and corrections
must  be  made, as required.   If  the  EICP area for  any  of  the  internal
standards changes by a factor of  two  (-50% to +100%) from the last  daily
calibration standard check,  the  mass  spectrometer must be inspected  for
malfunctions  and corrections must be made, as appropriate.

7.5   GC/MS  analysis

      7.5.1   It  is highly recommended  that  the  extract be screened on  a
GC/FID  or  GC/PID using the  same  type  of capillary column.   This  will

                             8270A - 13                        Revision 1
                                                                July 1992

-------
minimize  contamination  of  the  GC/MS  system  from  unexpectedly  high
concentrations of organic compounds.

      7.5.2   Spike  the  1  ml  extract  obtained from sample preparation  with
10 nl of the internal standard solution just prior to analysis.

      7.5.3   Analyze the  1 ml  extract by GC/MS  using  a  30 m x 0.25  mm (or
0.32 mm) silicone-coated fused-silica capillary column.   The volume to be
injected should ideally contain 100 ng of base/neutral  and 200 ng of  acid
surrogates  (for  a  1 /iL  injection).   The  recommended GC/MS  operating
conditions to be used are specified  in Section 7.3.

      7.5.4   If the response for  any quantitation ion exceeds the initial
calibration curve  range  of  the GC/MS system, extract dilution  must  take
place.  Additional  internal  standard must be added to the diluted extract
to  maintain  the  required   40  ng//iL  of each  internal   standard in  the
extracted volume.  The diluted extract must be reanalyzed.

      7.5.5   Perform all  qualitative  and   quantitative  measurements  as
described in Section 7.6.  Store the extracts at  4°C, protected from  light
in screw-cap vials equipped with unpierced Teflon lined septa.

7.6   Data  interpretation

      7.6.1   Qualitative  analysis

             7.6.1.1     An  analyte  (e.g.   those  listed in  Table  1)  is
      identified  by comparison of the sample mass  spectrum with the  mass
      spectrum of a standard of the suspected compound (standard reference
      spectrum).  Mass spectra for standard  reference  should be obtained on
      the isyr's  GC/MS within  the  same 12  hours as  the  sample analysis.
      These  standard reference  spectra may be obtained through analysis of
      the calibration standards.  Two criteria must be satisfied to  verify
      identification:   (1)  elution  of  sample  component  at the same  GC
      relative retention time  (RRT) as the standard component; and  (2)
      correspondence of the sample  component and the  standard  component
      mass  -.pectrum.

                   7.6.1.1.1   The sample component RRT must compare within
             + 0.06 RRT units  of the RRT of the  standard component.   For
             reference,  the  standard must  be run within the same 12 hours
             as  the sample.    If  coelution of  interfering  components
             prohibits accurate assignment of the sample component RRT from
             the  total ion  chromatogram,  the RRT  should  be  assigned  by
             using  extracted ion current profiles  for  ions unique  to the
             component of interest.

                   7.6.1.1.2   All   ions present   in  the  standard  mass
             spectrum  at  a relative  intensity  greater  than  10%  (most
             abundant ion in the spectrum  equals  100%  must be present in
             the  sample spectrum.

                   7.6.1.1.3   The relative intensities of ions specified
             in  Section  7.6.1.1.2 must  agree  within  plus or  minus  20%

                             8270A - 14                        Revision 1
                                                                July 1992

-------
       between the standard and  sample  spectra.   (Example:   For an
       ion with an  abundance  of 50% in  the  standard  spectrum,  the
       corresponding  sample  abundance  must  be  between  30  and  70
       percent.)

       7.6.1.2     For samples containing components not associated
with the calibration standards,  a library search  may be made for the
purpose of tentative identification.  The necessity to perform this
type of  identification' will  be  determined by  the purpose  of  the
analyses being conducted.  Computer generated library search routines
should not use normalization  routines that would  misrepresent  the
library or unknown spectra when compared to each other.  For example,
the  RCRA  permit or  waste  deli sting  requirements  may  require  the
reporting of  nontarget  analytes.  Only  after visual  comparison  of
sample  spectra with  the  nearest  library searches  will  the  mass
spectral interpretation specialist assign a tentative identification.
Guidelines for making tentative  identification are:

       (1) Relative  intensities  of major  ions  in the  reference
spectrum (ions > 10% of the most abundant ion) should be present in
the sample spectrum.

       (2) The relative intensities of  the-major ions  should agree
within ± 20%.   (Example:  For  an  ion with  an abundance of 50% in the
standard spectrum,  the  corresponding  sample  ion abundance  must  be
between 30 and 70%.)

       (3) Molecular ions present in the reference spectrum should be
present in the sample spectrum.

       (4) Ions  present  in  the sample  spectrum  but  not   in  the
reference  spectrum  should be   reviewed  for  possible  background
contamination or presence of coeluting compounds.

       (5) Ions present  in the  reference spectrum but not  in  the
sample spectrum should be reviewed  for possible subtraction from the
sample  spectrum  because of  background  contamination  or  coeluting
peaks.  Data system library reduction programs can sometimes create
these discrepancies.

7.6.2  Quantitative analysis

       7.6.2.1     When  a   compound   has  been   identified,   the
quantitation  of that  compound  will  be  based  on the  integrated
abundance  from  the   EICP  of  the  primary  characteristic  ion.
Quantitation will  take place using the internal  standard technique.
The  internal  standard  used shall be the  one  nearest  the retention
time of that of a given analyte  (e.g.  see Table 5).

       7.6.2.2     Calculate  the concentration   of  each identified
analyte in the sample as follows:
                       8270A - 15                        Revision 1
                                                          July 1992

-------
       Water:

       concentration (M9/L) =
                               (Ais)(RF)(Vo)(Vj)

where:

       Ax    =     Area  of  characteristic  ion  for  compound  being
                   measured.   •
       Is    =     Amount of internal standard injected (ng).
       Vt    =     Volume  of  total  extract,  taking  into  account
                   dilutions  (i.e.  a  l-to-10  dilution of  a  1  ml
                   extract will mean  Vt  =  10,000 /iL.   If half the
                   base/neutral extract  and half the  acid  extract
                   are combined,  Vt = 2,000).
       A.s    =     Area  of  characteristic  ion   for  the  internal
                   standard.
       RF    =     Response  factor  for  compound   being  measured
                   (Section 7.3.3).
       V0    =     Volume of water extracted (ml).
       V     =     Volume of extract injected
       Sediment/Soil  Sludge  (on  a  dry-weight  basis)  and  Waste
(normally on a wet-weight basis):
       concentration (/igAg) = —
                               (A

where:

       AX'  Is'  vt> Ais>  RF,  V. = Same as for water.
       Ws    =     Weight of sample extracted or diluted in grams.
       0     =     % dry weight of sample/100,  or  1 for a wet-weight
                   basis.

       7.6.2.3     Where appl icable, an estimate of concentration for
noncalibrated components in the sample should be made.  The formulas
given above  should  be  used with  the  following  modifications:   The
areas A  and A.  should be from the total ion chromatograms and the
RF for the  compound  should be assumed to be 1.   The concentration
obtained should  be reported  indicating  (1) that  the value'  is  an
estimate and  (2)  which internal  standard  was  used  to  determine
concentration.     Use   the  nearest   internal   standard   free  of
interferences.

       7.6.2.4     Quantitation  of multicomponent compounds   (e.g.
Aroclors)   is  beyond   the   scope  of  Method   8270A.    Normally,
quantitation is performed using a GC/ECD by Method 8080.
                       8270A - 16                        Revision 1
                                                          July 1992

-------
8.0   QUALITY  CONTROL

      8.1   Each laboratory  that  uses these  methods  is required to  operate a
formal quality control program. The minimum requirements of this program consist
of an initial  demonstration  of laboratory capability  and an ongoing analysis of
spiked  samples  to evaluate  and  document quality  data.   The  laboratory must
maintain records to document  the quality of  the  data generated.   Ongoing data
quality checks are compared with  established performance criteria to determine
if the results of analyses meet the performance characteristics of the method.
When  results  of sample  spikes  indicate  atypical  method performance,  a quality
control reference sample must be  analyzed to  confirm  that the measurements were
performed in  an in-control mode of operation.

      8.2   Before processing any  samples, the analyst must demonstrate, through
the analysis of a  reagent blank,  that interferences from the analytical system,
glassware,  and  reagents are  under  control.    Each time  a  set of  samples  is
extracted or there is  a change in reagents, a reagent blank should be processed
as a  safeguard against  chronic  laboratory  contamination.   The  blank samples
should be carried through all stages of sample preparation and measurement.

      8.3   The experience of the analyst performing GC/MS analyses is invaluable
to the success of the methods.  Each day that analysis is  performed, the daily
calibration standard  should be evaluated to determine  if  the chromatographic
system is operating  properly.   Questions  that should be asked are:  Do the peaks
look normal?;  Is the response obtained comparable to  the response from previous
calibrations?   Careful  examination  of the standard  chromatogram  can indicate
whether the column is still  good, the injector is leaking, the injector septum
needs  replacing,  etc.   If  any  changes  are  made  to the  system  (e.g.  column
changed), recalibration of the system must take place,

      8.4   Required instrument QC is found in the following sections

           8.4.1  The   GC/MS   system  must  be   tuned  to  meet   the   DFTPP
      specifications in  Sections  7.3.1 and 7.4.1.

           8.4,2  There must  be  an initial calibration  of  the GC/MS system as
      specified  in Section 7.3.

           8.4.3  The GC/MS  system  must  meet  the  SPCC criteria  specified  in
      Section  7.4.3  and  the  CCC criteria  in Section 7.4.4, each 12 hours.

      8.5   To  establish  the  ability   to generate  acceptable  accuracy  and
precision, the analyst must perform the  following operations.

           8.5.1  A  quality  control   (QC)   reference  sample  concentrate  is
      required containing each analyte at  a concentration of 100 mg/L in acetone.
      The  QC reference  sample  concentrate may  be prepared from  pure standard
      materials  or  purchased  as  certified  solutions.    If  prepared  by  the
      laboratory,  the  QC reference  sample concentrate must be  made using stock
      standards  prepared independently from those used  for calibration.

           8.5.2  Using a pipet,  prepare  QC reference samples at a concentration
      of 100 /ig/L  by adding  1.00  mL  of QC reference sample concentrate to each
      of four  1-L  aliquots of water.

                                  8270A  - 17                        Revision 1
                                                                     July 1992

-------
           8.5.3  Analyze the well-mixed QC reference samples according to the
     method beginning in Section 7.1. with extraction of the samples.

           8.5.4  Calculate the average recovery (x)  in  /xg/L, and the standard
     deviation  of the recovery (s)  in  /ig/L,  for each  analyte  using  the four
     results.

           8.5.5  For each  analyte  compare  s  and x with  the  corresponding
     acceptance  criteria_for  precision  and  accuracy,  respectively,  found  in
     Table 6.   If s and x for all analytes meet  the  acceptance criteria, the
     system performance is acceptable and analysis  of  actual  samples can_begin.
     If any individual s exceeds the precision limit or any individual x falls
     outside the range for accuracy, then the system performance is unacceptable
     for that analyte.

           NOTE:  The large number  of analytes  in Table 6 present a substantial
                  probability that one  or  more will  fail  at least  one  of the
                  acceptance criteria when all analytes of  a given  method are
                  analyzed.

           8.5.6  When one or more of the analytes tested  fail  at least one  of
     the  acceptance criteria,  the analyst must  proceed according  to Section
     8.5.6.1 or 8.5.6.2.

                  8.5.6.1     Locate  and correct the  source of  the problem and
           repeat the test for all  analytes beginning with  Section 8.5.2.

                  8.5.6.2     Beginning with Section 8.5.2,  repeat the  test only
           for those analytes that failed to  meet criteria.  Repeated failure,
           however, will  confirm a  general  problem with the measurement system.
           If ^his  occurs,  locate  and correct  the source  of the  problem and
           repeat the test  for all compounds  of interest  beginning with Step
           8.5.2.

     8.6   The laboratory must, on  an ongoing basis, analyze  a reagent blank, a
matrix spike,  end a replicate for each analytical  batch  (up to  a maximum of 20
samples/batch) to assess accuracy.  For  soil and waste samples where detectable
amounts of organics are present,  replicate samples may be  appropriate if' place
of matrix spiked  samples.   For laboratories  analyzing one  to ten  samples per
month,  at least one spiked sample per month is required.

           8.6.1  The concentration  of  the   spike in  the  sample  should  be
     determined as  follows:

                  8.6.1.1     If,  as  in compliance monitoring, the concentration
           of a  specific analyte  in the  sample  is  being  checked  against a
           regulatory concentration limit,  the  spike  should  be at that limit or
           1 to 5 times higher than the background concentration determined in
           Step 8.6.2, whichever concentration would be larger.

                  8.6.1.2     If the concentration of a specific analyte in a
           water  sample is not being  checked  against  a  limit specific to that
           analyte, the spike should  be at  100  jig/L or 1 to  5 times  higher than
           the  background  concentration determined  in  Step 8.6.2, whichever

                                  8270A - 18                        Revision 1
                                                                     July 1992

-------
           concentration would  be  larger.    For  other matrices,  recommended
           spiking concentration is 20 times the EQL.

                  8.6.1.3     If  it  is  impractical   to  determine  background
           levels before spiking (e.g. maximum holding times will be exceeded),
           the spike concentration  should be at (1) the regulatory concentration
           limit, if any; or, if none (2)  the larger of either 5 times higher
           than the expected background concentration or  100  ng/1.   For other
           matrices, recommended spiking concentration is  20 times the EQL.

           8.6.2  Analyze  one   sample  aliquot  to  determine  the  background
     concentration  (B)  of  each  analyte.    If necessary,  prepare  a new  QC
     reference  sample  concentrate  (Step 8.5.1)  appropriate  for the background
     concentration  in  the sample.   Spike a second  sample  aliquot with 1.00  ml
     of  the  QC reference sample concentrate  and  analyze  it  to  determine  the
     concentration  after spiking  (A)  of each  analyte.  Calculate each percent
     recovery  (p) as 100(A-B)%/T, where  T  is the known true value of the spike.

           8.6.3  Compare the percent recovery (p)  for each  analyte in a water
     sample  with the corresponding QC  acceptance  criteria  found  in  Table  6.
     These acceptance criteria were calculated to include an allowance for error
     in  measurement of both  the background and  spike concentrations,  assuming
     a spike  to background ratio  of 5:1.   This  error will  be  accounted for  to
     the extent that the analyst's spike  to  background ratio approaches 5:1.
     If  spiking was  performed  at  a  concentration  lower than 100 fig/L,  the
     analyst must use  either the  QC acceptance  criteria presented in  Table  6,
     or  optional QC  acceptance criteria  calculated for  the  specific  spike
     concentration.  To calculate optional  acceptance criteria  for the recovery
     of  an  analyte: (1) Calculate  accuracy (x')  using the equation  found  in
     Table  7,  substituting  the spike concentration  (T)  for  C;  (2)  calculate
     overall precision  (S') using  the equation  in Table 7, substituting x'  for
     x;  (3)  calculate  the range  for recovery  at  the spike  concentration  as
     (lOOx'/T)  ± 2.44(100S7T)%.

           8.6.4  If any individual  p falls outside the  designated  range  for
     recovery,  that  analyte  has  failed   the  acceptance  criteria.    A  check
     standard containing each analyte that  failed the criteria  must be analyzed
     as  described in Section 8.7.

     8.7  If any analyte in a  water  sample fails  the  acceptance criteria  for
recovery  in  Section 8.6,  a QC  reference  sample containing each  analyte that
failed must be prepared and analyzed.

     NOTE;        The  frequency  for the required analysis  of a  QC  reference
                  sample  will   depend  upon  the  number   of   analytes  being
                  simultaneously tested, the  complexity of  the sample matrix,
                  and the performance of the laboratory.   If the entire list of
                  analytes in Table 6  must  be  measured in the sample in Section
                  8.6,  the  probability  that  the  analysis  of a  QC  reference
                  sample will   be  required is  high.   In  this  case the  QC
                  reference sample should be routinely analyzed with the spiked
                  sample.
                                  8270A - 19                        Revision 1
                                                                     July 1992

-------
           8.7.1  Prepare the QC  reference  sample by adding 1.0 ml  of the QC
     reference sample concentrate (Section 8.5.1 or 8.6.2) to 1  L of water.  The
     QC reference sample needs only to contain the  analytes that  failed criteria
     in the test  in Section 8.6.

           8.7.2  Analyze the QC reference sample to determine the concentration
     measured  (A) of each analyte.   Calculate each  percent  recovery (ps) as
     100(A/T)%, where T  is the true value of  the  standard concentration.

           8.7.3  Compare the percent recovery  (ps)  for  each analyte with the
     corresponding QC acceptance criteria found in Table  6.  Only analytes that
     failed the test  in  Step 8.6  need  to be compared with these criteria.  If
     the recovery of  any such analyte  falls outside  the  designated  range, the
     laboratory performance for that analyte is judged to  be  out  of control, and
     the problem  must be  immediately identified and corrected.   The  analytical
     result for that  analyte in the unspiked  sample  is suspect  and  may not be
     reported for regulatory compliance  purposes.

     8.8   As part of the QC program for the laboratory, method accuracy for each
matrix studied  must  be  assessed  and  records must be maintained.    After the
analysis of five spiked samples  (of the same matrix) as in  Section 8.6, calculate
the average  percent  recovery (p)  and  the  standard  deviation   of the percent
recovery (s ).   Express  the accuracy  assessment as a  percent recovery interval
from p - 2s  to p +  2s  .   If p  = 90% and s   = 10%,  for  example,  the accuracy
interval  is  expressed as  70-110%.   Update the  accuracy assessment  for each
analyte  on  a   regular   basis  (e.g.   after  each   five   to  ten new  accuracy
measurements).

     8.9   To determine acceptable accuracy and precision limits for surrogate
standards the following procedure should be performed.

           8.9.1  For each sample  analyzed, calculate the percent  recovery of
     each  surrogate  in the sample.

           8.9.2  Once a minimum of thirty samples of the  same matrix have been
     analyzed, calculate the average percent recovery (p)  and standard deviation
     of the percent recovery (s)  for each of  the  surrogates.

           8.9.3  For a  given  matrix,  calculate  the upper and lower control
     limit for method performance for each  surrogate  standard.   This should be
     done  as follows:

           Upper  Control Limit  (UCL)  = p + 3s
           Lower  Control Limit  (LCL)  = p - 3s

           8.9.4  For aqueous and  soil  matrices,  these  laboratory established
     surrogate  control   limits  should,  if  applicable,   be  compared  with the
     control limits  listed  in Table 8.   The limits given  in Table 8 are multi-
     laboratory  performance based limits for  soil  and   aqueous  samples, and
     therefore, the single-laboratory limits established  in Step 8.9.3 must fall
     within those given  in Table  8 for  these  matrices.
                                  8270A - 20                        Revision 1
                                                                     July 1992

-------
           8.9.5  If recovery is not within limits, the following procedures are
     required.

           •      Check  to  be  sure  there  are  no  errors  in  calculations,
                  surrogate  solutions  and  internal  standards.   Also,  check
                  instrument performance.

           •      Recalculate the data and/or  reanalyze  the  extract  if any of
                  the above checks reveal a problem.

           •      Reextract and reanalyze the sample if none  of the above are a
                  problem or flag the data as "estimated concentration".

           8.9.6  At a minimum, each laboratory should update  surrogate recovery
     limits on  a matrix-by-matrix basis,  annually.

     8.10  It  is  recommended  that  the  laboratory  adopt additional  quality
assurance practices for use with this method.  The specific practices that are
most productive depend upon the needs of the  laboratory  and  the  nature of the
samples.   Field duplicates  may  be analyzed  to  assess  the  precision  of the
environmental  measurements.  When doubt exists over the identification of a peak
on the chromatogram, confirmatory techniques such as gas chromatography with a
dissimilar column,  specific element detector, or mass spectrometer must be used.
Whenever possible, the laboratory should analyze standard reference materials and
participate in relevant performance evaluation studies.


9.0  METHOD PERFORMANCE

     9.1   Method 8250 (the packed column version  of Method 8270) was tested by
15 laboratories using Organic-free reagent water, drinking water, surface water,
and  industrial  wastewaters  spiked  at six  concentrations over  the  range  5-
1,300 M9/L.   Single operator  accuracy and precision,  and method  accuracy were
found to be directly related to the concentration of the analyte and essentially
independent  of  the   sample  matrix.  Linear   equations  to  describe  these
relationships are presented in Table 7. Method performance data for Method 8270
is being developed.


10.0 REFERENCES

1.   U.S.  EPA 40 CFR Part 136,  "Guidelines Establishing Test  Procedures for the
     Analysis of Pollutants Under the Clean Water Act,  Method 625," October 26,
     1984.

2.   U.S.  EPA  Contract   Laboratory  Program,  Statement   of  Work for  Organic
     Analysis,  July  1985,  Revision.

3.   Eichelberger,  J.W.,  L,E.  Harris,  and W.L.  Budde,  "Reference  Compound to
     Calibrate  Ion Abundance Measurement  in Gas Chromatography-Mass Spectrometry
     Systems,"  Analytical  Chemistry, 47,  995-1000,  1975.

4.   "Method  Detection Limit for Methods  624 and  625," Olynyk, P.,  W.L. Budde,
     and J.W.  Eichelberger,  Unpublished  report,  October  1980.

                                  8270A  - 21                        Revision  1
                                                                     July 1992

-------
5.     "Inter-laboratory Method Study for EPA Method 625-Base/Neutrals, Acids, and
      Pesticides,"  Final  Report  for  EPA  Contract  68-03-3102  (in preparation).

6.     Burke,  J.A.  "Gas  Chromatography  for  Pesticide  Residue  Analysis;  Some
      Practical  Aspects," Journal  of the  Association  of  Official  Analytical
      Chemists,  48,  1037,  1965.

7.     Lucas,  S.V.;  Kornfeld, R.A. "GC-MS  Suitability Testing of RCRA Appendix
      VIII  and  Michigan  List Analytes "; U.S. Environmental  Protection Agency,
      Environmental  Monitoring  and  Support Laboratory, Cincinnati,  OH 45268,
      February  20,  1987,  Contract No.  68-03-3224.

8.     Engel,  T.M.;  Kornfeld, R.A.; Warner, J.S.;  Andrews,  K.D.  "Screening of
      Semivolatile  Organic Compounds for Extractabllity  and  Aqueous Stability by
      SW-846,  Method 3510"; U.S. Environmental  Protection Agency, Environmental
      Monitoring and Support  Laboratory,  Cincinnati,  OH 45268,  June  5,  1987,
      Contract  68-03-3224.
                                   8270A -  22                         Revision 1
                                                                      July 1992

-------
                   TABLE 1.
CHARACTERISTIC IONS FOR SEMIVOLATILE COMPOUNDS
Compound
2-Picoline
Aniline
Phenol
Bis(2-chloroethyl) ether
2-Chlorophenol
1,3-Dichlorobenzene
l,4-Dichlorobenzene-d4 (I.S.)
1,4-Dichlorobenzene
Benzyl alcohol
1,2-Dichlorobenzene
N-Ni trosomethyl ethyl ami ne
Bis(2-chloroisopropyl) ether
Ethyl carbamate
Thiophenol (Benzenethiol )
Methyl methanesulfonate
N-Nitrosodi -n-propylamine
Hexachloroethane
Maleic anhydride
Nitrobenzene
Isophorone
N-Nitrosodi ethyl ami ne
2-Nitrophenol
2,4-Dimethylphenol
p-Benzoquinone
Bis(2-chloroethoxy)methane
Benzoic acid
2,4-Dichlorophenol
Trimethyl phosphate
Ethyl methanesulfonate
1,2,4-Trichlorobenzene
Naphthalene-da (I.S.)
Naphthalene
Hexachl orobutadi ene
Tetraethyl pyrophosphate
Diethyl sulfate
4-Chl oro-3-methyl phenol
2-Methyl naphthalene
2-Methyl phenol
Hexachl oropropene
Hexachl orocycl opentadi ene
N-Nitrosopyrrol idine
Acetophenone
4-Methyl phenol
2,4,6-Trichlorophenol
o-Toluidine
3-Methylphenol
2-Chloronaphthalene
Retention
Time (min.)
3.75a
5.68
5.77
5.82
5.97
6.27
6.35
6.40
6.78
6.85
6.97
7.22
7.27
7.42
7.48
7.55
7.65
7.65
7.87
8.53
8.70
8.75
9.03
9.13
9.23
9.38
9.48
9.53
9.62
9.67
9.75
9.82
10.43
11.07
11.37
11.68
11-. 87
12.40
12.45
12.60
12.65
12.67
12.82
12.85
12.87
12.93
13.30
Primary
Ion
93
93
94
93
128
146
152
146
108
146
88
45
62
110
80
70
117
54
77
82
102
139
122
108
93
122
162
110
79
180
136
128
225
99
139
107
142
107
213
237
100
105
107
196
106
107
162
Secondary
Ion(s)
66,92
66,65
65,66
63,95
64,130
148,111
150,115
148,111
79,77
148,111
42,88,43,56
77,121
62,44,45,74
110,66,109,84
80,79,65,95
42,101,130
201,199
54,98,53,44
123,65
95,138
102,42,57,44,56
109,65
107,121
54,108,82,80
95,123
105,77
164,98
110,79,95,109,140
79,109,97,45,65
182,145
68
129,127
223,227
99,155,127,81,109
139,45,59,99,111,125
144,142
141
107,108,77,79,90
213,211,215,117,106,141
235,272
100,41,42,68,69
71,105,51,120
107,108,77,79,90
198,200
106,107,77,51,79
107,108,77,79,90
127,164
                  8270A  -  23
Revision 1
 July 1992

-------
                                   TABLE 1.
                                  (Continued)
Compound
Retention
Time (min.)
Primary
Ion
Secondary
Ion(s)
N-Nitrosopiperidine                    13.55
1,4-Phenylenediamine                   13.62
1-Chloronaphthalene                    13.658
2-Nitroaniline                         13.75
5-Ch1oro-2-methylani1ine               14.28
Dimethyl phthalate                     14.48
Acenaphthylene                         14.57
2,6-Dinitrotoluene                     14.62
Phthalic anhydride                     14.62
o-Anisidine                            15.00
3-Nitroaniline                         15.02
Acenaphthene-d10  (I.S.)                 15.05
Acenaphthene                           15.13
2,4-Dinitrophenol                      15.35
2,6-Dinitrophenol                      15.47
4-Chloroaniline                        15.50
Isosafrole                             15.60
Dibenzofuran                           15.63
2,4-Diaminotoluene                     15.78
2,4-Dinitrotoluene                     15.80
4-Nitrophenol                          15.80
2-Naphthylamine                        16.00a
1,4-Naphthoquinone                     16.23
p-Cresidine                            16.45
Oichlorovos                            16.48
Diethyl phthalate                      16.70
Fluorene                               16.70
2,4,5-Trimethylaniline                 16.70
N-Nitrosodibutylamine                  16.73
4-Chlorophenyl phenyl ether            16.78
Hydroquinone                           16.93
4,6-Dinitro-2-methylpheno1             17.05
Resorcinol                             17.13
N-Nitrosodiphenylamine                 17.17
Safrole                                17.23
Hexamethyl phosphoramide               17.33
3-(Chloromethy1)pyridine hydrochloride!7,50
Diphenylamine                          17.548
1,2,4,5-Tetrachlorobenzene             17.97
1-Naphthylamine                        18.20
l-Acetyl-2-thiourea                    18.22
4-Bromophenyl phenyl ether             18.27
Toluene diisocyanate                   18.42
2,4,5-Trichlorophenol                  18.47
Hexachlorobenzene                      18.65
Nicotine                               18.70
Pentachlorophenol                      19.25

                                  8270A  - 24
114     42,114,55,56,41
108     108,80,53,54,52
162     127,164
 65     92,138
106     106,141,140,77,89
163     194,164
152     151,153
165     63,89
104     104,76,50,148
108     80,108,123,52
138     108,92
164     162,160
154     153,152
184     63,154
162     162,164,126,98,63
127     127,129,65,92
162     162,131,104,77,51
168     139
121     121,122,94,77,104
165     63,89
139     109,65
143     115,116
158     158,104,102,76,50,130
122     122,94,137,77,93
109     109,185,79,145
149     177,150
166     165,167
120     120,135,134,91,77
 84     84,57,41,116,158
204     206,141
110     110,81,53,55
198     51,105
110     110,81,82,53,69
169     168,167
162     162,162,104,77,103,135
135     135,44,179,92,42
 92     92,127,129,65,39
169     168,167
216     216,214,179,108,143,218
143     143,115,89,63
118     43,118,42,76
248     250,141
174     174,145,173,146,132,91
196     196,198,97,132,99
284     142,249
 84     84,133,161,162
266     264,268

                    Revision 1
                     July  1992

-------
                                   TABLE 1.
                                  (Continued)
Compound
Retention
Time (min.)
Primary   Secondary
Ion       Ion(s)
5-Nitro-o-toluidine
Thionazine
4-Nitroaniline
Phenanthrene-d10( i. s.)
Phenanthrene
Anthracene
1,4-Dinitrobenzene
Mevinphos
Naled
1,3-Oinitrobenzene
Diallate (cis or trans)
1,2-Dinitrobenzene
Diallate (trans or cis)
Pentachlorobenzene
5-Nitro-o-anisidine
Pentachloronitrobenzene
4-Nitroquinoline-1-oxide
Di-n-butyl phthalate
2,3,4,6-Tetrachlorophenol
Demeton-0
Fluoranthene
1,3,5-Trinitrobenzene
Dicrotophos
Benzidine
Trifluralin
Bromoxynil
Pyrene
Monocrotophos
Phorate
Sulfall ate
Demeton-S
Phenacetin
Dimethoate
Phenobarbital
Carbofuran
Octamethyl pyrophosphoramide
4-Aminobiphenyl
Terbufos
a,a-Oimethylphenylamine
Pronamide
Aminoazobenzene
Dichlone
Dinoseb
Disulfoton
Fluchloralin
Mexacarbate
4,4'-Oxydianiline
    19.27      152     77,152,79,106,94
    19.35      107     96,107,97,143,79,68
    19.37      138     138,65,108,92,80,39
    19.55      188     94,80
    19.62      178     179,176
    19.77      178     176,179
    19.83      168     168,75,50,76,92,122
    19.90      127     127,192,109,67,164
    20.03      109     109,145,147,301,79,189
    20.18      168     168,76,50,75,92,122
    20.57      86     86,234,43,70
    20.58      168     168,50,63,74
    20.78      86     86,234,43,70
    21.35      250     250,252,108,248,215,254
    21.50      168     168,79,52,138,153,77
    21.72      237     237,142,214,249,295,265
    21.73      174     174,101,128,75,116
    21.78      149     150,104
    21.88      232     232,131,230,166,234,168
    22.72      88     88,89,60,61,115,171
    23.33      202     101,203
    23.68      75     75,74,213,120,91,63
    23.82      127     127,67,72,109,193,237
    23.87      184     92,185
    23.88      306     306,43,264,41,290
    23.90      277     277,279,88,275,168
    24.02      202     200,203
    24.08      127     127,192,67,97,109
    24.10      75     75,121,97,93,260
    24.23      188     188,88,72,60,44
    24.30      88     88,60,81,89,114,115
    24.33      108     180,179,109,137,80
    24.70      87     87,93,125,143,229
    24.70      204     204,117,232,146,161
    24.90      164     164,149,131,122
    24.95      135     135,44,199,286,153,243
    25.08      169     169,168,170,115
    25.35      231     231,57,97,153,103
    25.43      58     58,91,65,134,42
    25.48      173     173,175,145,109,147
    25.72      197     92,197,120,65,77
    25.77      191     191,163,226,228,135,193
    25.83      211     211,163,147,117,240
    25.83      88     88,97,89,142,186
    25.88      306     306,63,326,328,264,65
    26.02      165     165,150,134,164,222
    26.08      200     200,108,171,80.65

8270A - 25                        Revision  1
                                   July  1992

-------
                                   TABLE 1.
                                  (Continued)
Compound
Retention
Time (min.)
Primary   Secondary
Ion       Ion(s)
Butyl benzyl phthalate
4-Nitrobiphenyl
Phosphamidon
2-Cyclohexyl-4,6-Dinitrophenol
Methyl parathion
Carbaryl
Dimethyl aminoazobenzene
Propylthiouracil
Benz(a)anthracene
Chrysene-d12 (I.S.)
3,3'-Dichlorobenzidine
Chrysene
Malathion
Kepone
Fenthion
Parathion
Anilazine
Bis(2-ethylhexyl) phthalate
3,3'-Dimethylbenzidine
Carbophenothion
5-Nitroacenaphthene
Methapyrilene
Isodrin
Captan
Chlorfenvinphos
Crotoxyphos
Phosmet
EPN
Tetrachlorvinphos
Di-n-octyl  phthalate
2-Ami noanthraqu i none
Barban
Aramite
Benzo(b)f1uoranthene
Nitrofen
Benzo(k)fl uoranthene
Chiorobenzilate
Fensulfothion
Ethion
Diethylstilbestrol
Famphur
Tri-p-tolyl phosphate
Benzo(a)pyrene
Perylene-d)2 (I.S.)
7,12-Dimetnylbenz(a)anthracene
5,5-Di phenylhydantoi n
Captafol
    26.43     149     91,206
    26.55     199     199,152,141,169,151
    26.85     127     127,264,72,109,138
    26.87     231     231,185,41,193,266
    27.03     109     109,125,263,79,93
    27.17     144     144,115,116,201
    27.50     225     225,120,77,105,148,42
    27.68     170     170,142,114,83
    27.83     228     229,226
    27.88     240     120,236
    27.88     252     254,126
    27.97     228     226,229
    28.08     173     173,125,127,93,158
    28.18     272     272,274,237,178,143,270
    28.37     278     278,125,109,169,153
    28.40     109     109,97,291,139,155
    28.47     239     239,241,143,178,89
    28.47     149     167,279
    28.55     212     212,106,196,180
    28.58     157     157,97,121,342,159,199
    28.73     199     199,152,169,141,115
    28.77      97     97,50,191,71
    28.95     193     193,66,195,263,265,147
    29.47      79     79,149,77,119,117
    29.53     267     267,269,323,325.295
    29.73     127     127,105,193,166
    30.03     160     160,77,93,317,76
    30.11     157     157,169,185,141,323
    30.27     329     109,329,331,79,333
    30.48     149     167,43
    30.63     223     223,167,195
    30.83     222     222,51,87,224,257,153
    30.92     185     185,191,319,334,197,321
    31.45     252     253,125
    31.48     283     283,285,202,139,253
    31.55     252     253,125
    31.77     251     251,139,253,111,141
    31.87     293     293,97,308,125,292
    32.08     231     231,97,153,125,121
    32.15     268     268,145,107,239,121,159
    32.67     218     218,125,93,109,217
    32.75     368     358,367,107,165,198
    32.80     252     253,125
    33.05     264     260,265
    33.25     256     256,241,239,120
    33.40     180     180,104,252,223,209
    33.47      79     79,77,80,107
                                   8270A -  26
                                  Revision 1
                                   July 1992

-------
                                   TABLE 1.
                                  (Continued)
Compound
Retention
Time (min.)
Primary   Secondary
Ion       Ion(s)
Dinocap
Methoxychlor
2-Acetylami nof1uorene
4,4'-Methylenebis(2-ch1oroaniline)
3,3'-Dimethoxybenzidine
3-Methylcholanthrene
Phosalone
Azinphos-methyl
Leptophos
Mirex
Tris(2,3-dibromopropyl) phosphate
Dibenz(a,j)acridine
Mestranol
Coumaphos
Indeno(l,2,3-cd)pyrene
Dibenz(a,h)anthracene
Benzo(g,h,i)perylene
l,2:4,5-Dibenzopyrene
Strychnine
Piperonyl sulfoxide
Hexachlorophene
Aldrin
Aroclor-1016
Aroclor-1221
Aroclor-1232
Aroclor-1242
Aroclor-1248
Aroclor-1254
Aroclor-1260
a-BHC
0-BHC
5-BHC
Y-BHC  (Lindane)
4,4'-DDD
4,4'-DDE
4,4/-ODT
Dieldrin
1,2-Diphenylhydrazine
Endosulfan I
Endosulfan II
Endosulfan sulfate
Endrin
Endrin aldehyde
Endrin ketone
2-Fluorobiphenyl (surr.)
2-Fluorophenol (surr.)
    33.47      69     69,41,39
    33.55     227     227,228,152,114,274,212
    33.58     181     181,180,223,152
    34.38     231     231,266,268,140,195
    34.47     244     244,201,229
    35.07     268     268,252,253,126,134,113
    35.23     182     182,184,367,121,379
    35.25     160     160,132,93,104,105
    35.28     171     171,377,375,77,155,379
    35.43     272     272,237,274,270,239,235
    35.68     201     137,201,119,217,219,199
    36.40     279     279,280,277,250
    36.48     277     277,310,174,147,242
    37.08     362     362,226,210,364,97,109
    39.52     276     138,227
    39.82     278     139,279
    41.43     276     138,277
    41.60     302     302,151,150,300
    45.15     334     334,335,333
    46.43     162     162,135,105,77
    47.98     196     196,198,209,211,406,408
               66     263,220
              222     260,292
              190     224,260
              190     224,260
              222     256,292
              292     362,326
              292     362,326
              360     362,394
              183     181,109
              181     183,109
              183     181,109
              183     181,109
              235     237,165
              246     248,176
              235     237,165
               79     263,279
               77     105,182
              195     339,341
              337     339,341
              272     387,422
              263     82,81
               67     345,250
              317     67,319
              172     171
              112     64
                                  8270A - 27
                                  Revision 1
                                   July 1992

-------
                                   TABLE 1.
                                  (Continued)
                                  Retention     Primary    Secondary
Compound                          Time (min.)   Ion        Ion(s)
Heptachlor                             --        100     272,274
Heptachlor epoxide                     --        353     355,351
Nitrobenzene-d5 (surr.)                 --         82     128,54
N-Nitrosodimethylamine                 --         42     74,44
Phenol-d, (surr.)                       --         99     42,71
Terphenyl-du  (surr.)                   --        244     122,212
2,4,6-Tribromophenol (surr.)           --        330     332,141
Toxaphene                              --        159     231,233
I.S. = internal standard.
surr. = surrogate.
8Estimated retention times.
Substitute for the non-specific mixture,  tricresyl  phosphate.
                                  8270A  -  28                        Revision  1
                                                                     July  1992

-------
                           TABLE 2.
ESTIMATED QUANTITATION LIMITS (EQLs) FOR SEMIVOLATILE ORGANICS8
Semivolatiles
Acenaphthene
Acenaphthylene
Acetophenone
2-Acetylaminofluorene
l-Acetyl-2-thiourea
2-Aminoanthraquinone
Aminoazobenzene
4-Anrinobiphenyl
Anilazine
o-Anisidine
Anthracene
Aramite
Azinphos-methyl
Barban
Benz( a) anthracene
Benzo(b)fluoranthene
Benzo ( k) fl uoranthene
Benzole acid
Benzo(g,h,i)perylene
Benzo(a)pyrene
p-Benzoquinone
Benzyl alcohol
Bis(2-chloroethoxy)methane
Bis(2-chloroethyl ) ether
Bis(2-chloroisopropyl) ether
4-bromophenyl phenyl ether
Bromoxynil
Butyl benzyl phthalate
Captafol
Captan
Carbaryl
Carbofuran
Carbophenothion
Chlorfenvinphos
4-ChToroaniline
Chlorobenzilate
5-Chloro-2-methyl aniline
4-Chloro-3-methyl phenol
3-(Chloroniethyl)pyridine hydrochloride
2-Chloronaphthalene
2-Chlorophenol
4-Chlorophenyl phenyl ether
Chrysene
Coumaphos
Estimated
Quantitation
Limits6
Ground water Low Soi
M9/L
10
10
10
20
1000
20
10
20
100
10
10
20
100
200
10
10
10
50
10
10
10
20
10
10
10
10
10
10
20
50
10
10
10
20
20
10
10
20
100
10
10
10
10
40

I/Sediment'
M9/kg
660
660
NO
NO
ND
ND
ND
ND
ND
NO
660
ND
ND
ND
660
660
660
3300
660
660
ND
1300
660
660
660
660
ND
660
ND
ND
ND
ND
ND
ND
1300
ND
ND
1300
ND
660
660
660
660
ND
                          8270A - 29                        Revision 1
                                                             July 1992

-------
TABLE 2.
(Continued)
Semivolatiles
p-Cresidine
Crotoxyphos
2-Cyclohexyl -4,6-dinitrophenol
Demeton-0
Demeton-S
Diallate (cis or trans)
Diallate (trans or cis)
2,4-Diaminotoluene
Dibenz(a,j)acridine
Dibenz( a, h) anthracene
Dibenzofuran
Dibenzo(a,e)pyrene
Di-n-butyl phthalate
Dichlone
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
3,3'-Oichlorobenzidine
2,4-Dichlorophenol
2,6-Dichlorophenol
Dichlorovos
Dicrotophos
Diethyl phthalate
Diethylstilbestrol
Diethyl sulfate
Dimethoate
3,3'-Dimethoxybenzidine
Dimethyl aminoazobenzene
7, 12 -Dimethyl benz( a) anthracene
3,3'-Oimethylbenzidine
a , a-Di methyl phenethyl ami ne
2,4-Dimethylphenol
Dimethyl phthalate
1,2-Dinitrobenzene
1,3-Di nitrobenzene
1,4-Dinitrobenzene
4, 6-Dinitro-2-methyl phenol
2,4-D'initrophenol
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Dinocap
Dinoseb
5,5-Oiphenylhydantoin
Di-n-octyl phthalate
Estimated
Quantitation
Limits"
Ground water Low Soi
M9/L
10
20
100
10
10
10
10
20
10
10
10
10
10
NA
10
10
10
20
10
10
10
10
10
20
100
20
100
10
10
10
ND
10
10
40
20
40
50
50
10
10
100
20
20
10

1 /Sediment'
M9/kg
ND
ND
ND
ND
ND
ND
ND
ND
ND
660
660
ND
ND
ND
660
660
660
1300
660
ND
ND
ND
660
ND
ND
ND
ND
ND
ND
ND
ND
660
660
ND
ND
ND
3300
3300
660
660
NO
ND
ND
660
8270A - 30
Revision 1
 July 1992

-------






Semivolatiles
Disulfoton
EPN
Ethion
Ethyl carbamate
Bis(2-ethylhexyl) phthalate
Ethyl methanesulfonate
Famphur
Fensulfothion
Fenthion
Fluchloral in
Fluoranthene
Fluorene
Hexachl orobenzene
Hexachlorobutadiene
Hexachl orocyclopentadiene
Hexachloroethane
Hexachlorophene
Hexachl oropropene
Hexamethylphosphoramide
Hydroquinone
Indeno(l,2,3-cd)pyrene
Isodrin
Isophorone
Isosafrole
Kepone
Leptophos
Malathion
Maleic anhydride
Mestranol
Me.thapyrilene
Methoxychlor
3-Methyl chol anthrene
4,4' -Methylenebi.s(2-chloroani
Methyl methanesulfonate
2-Methylnaphthalene
Methyl parathion
2-Methyl phenol
3-Methyl phenol
4-Methyl phenol
Mevinphos
Mexacarbate
Mirex
Monocrotophos
Naled
TABLE 2.
(Continued)
Estimated
Quantisation
. Limits
Ground water Low Soi
M9/L
10
10
10
50
10
20
20
40
10
20
10
10
10
10
10
10
50
10
20
ND
10
20
10
10
20
10
50
NA
20
100
10
10
line) NA
10
10
10
10
10
10
10
20
10
40
20





I/Sediment1
MQAg
NO
ND
ND
ND
660
ND
ND
ND
ND
ND
660
660
660
660
660
660
ND
ND
ND
ND
660
ND
660
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
660
ND
660
ND
660
ND
ND
ND
ND
ND
8270A - 31
Revision 1
 July 1992

-------






Semivolatiles
Naphthalene
1,4-Naphthoquinone
1-Naphthylamine
2-Naphthylamine
Nicotine
5-Nitroacenaphthene
2-Nitroaniline
3-Nitroaniline
4-Nitroaniline
5-Nitro-o-anisidine
Nitrobenzene
4-Nitrobiphenyl
Nitrofen
2-Nitrophenol
4-Nitrophenol
5-Nitro-o-toluidine
4-Nitroquinoline-l-oxide
N-Ni trosodi butyl ami ne
N-Ni trosodi ethyl ami ne
N-Ni trosodi phenyl ami ne
N-Ni troso-di-n-propyl ami ne
N-Nitrosopiperidine
N-Nitrosopyrrol idine
Octamethyl pyrophosphoramide
4,4'-Oxydianiline
Parathion
Pentachlorobenzene
Pentachloronitrobenzene
Pentachlorophenol
Phenacetin
Phenanthrene
Phenobarbital
Phenol
1,4- Phenyl enedi ami ne
Phorate
Phosalone
Phosmet
Phosphamidon
Phthalic anhydride
2-Picoline
FMperonyl sulfoxide
Pronamide
Propylthiouracil
Pyrene
TABLE 2.
(Continued)
Estimated
Quantitation
Limits"
Ground water Low Soi
M9/L
10
10
10
10
20
10
50
50
20
10
10
10
20
10
50
10
40
10
20
10
10
20
40
200
20
10
10
20
50
20
10
10
10
10
10
100
40
100
100
NO
100
10
100
10





I/Sediment1
M9/kg
660
NO
NO
NO
NO
ND
3300
3300
ND
ND
660
ND
ND
660
3300
ND
ND
ND
ND
660
660
ND
ND
ND
ND
ND
ND
ND
3300
ND
660
ND
660
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
660
8270A - 32
Revision 1
 July 1992

-------
                                   TABLE 2.
                                  (Continued)
                                               Estimated
                                              Quantitation
                                                 Limits"
                                      Ground  water    Low  Soil/SedimenT
Semivolatiles                            M9/L
Pyridine
Resorcinol
Safrole
Strychnine
Sul fall ate
Terbufos
1,2,4 , 5-Tetrachl orobenzene
2,3,4,6-Tetrachlorophenol
Tetrachlorvinphos
Tetraethyl pyrophosphate
Thionazine
Thiophenol (Benzenethiol)
Toluene diisocyanate
o-Toluidine
1, 2, 4-Trichl orobenzene
2,4,5-Trichlorophenol
2,4,6-Trichlorophenol
Trifluralin
. 2, 4, 5-Tri methyl aniline
Trimethyl phosphate
1,3, 5-Tri nitrobenzene
Tris(2,3-dibromopropyl ) phosphate
Tri-p-tolyl phosphate(h)
0,0,0-Tri ethyl phosphorothioate
ND
100
10
40
10
20
10
10
20
40
20
20
100
10
10
10
10
10
10
10
10
200
10
NT
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
660
660
660
ND
ND
ND
ND
ND
ND
ND
a EQLs  listed  for soil/sediment  are  based on wet  weight.   Normally  data  is
  reported on a dry weight  basis,  therefore,  EQLs will  be  higher based on the
  %  dry weight  of each  sample.    This  is based  on a  30 g  sample  and  gel
  permeation chromatography cleanup.
b Sample EQLs are highly matrix-dependent.   The EQLs listed herein are provided
  for guidance and may not always be achievable.
ND = Not determined.
NA = Not applicable.
NT = Not tested.
Other Matrices                                       Factor1

High-concentration soil and sludges by ultrasonic extractor  7.5
Non-water miscible waste                                    75

1EQL = [EQL for Low Soil/Sediment (Table 2)]  X [Factor].
                                  8270A - 33                        Revision 1
                                                                     July 1992

-------
                                   TABLE  3.
                  DFTPP KEY IONS AND ION  ABUNDANCE CRITERIA8
       Mass
Ion Abundance Criteria
       51

       68
       70

      127

      197
      198
      199

      275

      365

      441
      442
      443
30-60% of mass 198

< 2% of mass 69
< 2% of mass 69

40-60% of mass 198

< 1% of mass 198
Base peak, 100% relative abundance
5-9% of mass 198

10-30% of mass 198

> 1% of mass 198

Present but less than mass 443
> 40% of mass 198
17-23% of mass 442
   aSee  Reference  4.
                                   TABLE 4.
                          CALIBRATION CHECK COMPOUNDS
Base/Neutral Fraction
                  Acid Fraction
Acenaphthene
1,4-Dichlorobenzene
Hexachlorobutadiene
N-Nitrosodiphenylamine
Di-n-octyl phthalate
Fluoranthene
Benzo(a)pyrene
                  4-Chloro-3-methyl phenol
                  2,4-Dichlorophenol
                  2-Nitrophenol
                  Phenol
                  Pentachlorophenol
                  2,4,6-Trichlorophenol
                                  8270A - 34
                                            Revision 1
                                             July 1992

-------
                                   TABLE 5.
          SEMIVOLATILE  INTERNAL STANDARDS WITH CORRESPONDING ANALYTES
                           ASSIGNED FOR QUANTITATION
!,4-Dich1orobenzene-d4
Naphthalene-d.
Acenaphthene-d
                                            10
Aniline
Benzyl alcohol
Bis(2-chloroethyl) ether
Bis(2-chloroisopropyl) ether
2-Chlorophenol
1,3-Dichlorobenzene
1,4-Dichlorobenzene
1,2-Dichlorobenzene
Ethyl methanesulfonate
2-Fluorophenol (surr.)
Hexachloroethane
Methyl methanesulfonate
2-Methylphenol
4-Methylphenol
N-Nitrosodimethylamine
N-Nitroso-di-n-propylamine
Phenol
Phenol-d6 (surr.)
2-Picoline
Acetophenone
Benzoic acid
Bis(2-chloroethoxy)methane
4-Chloroaniline
4-Chloro-3-methylphenol
2,4-Dichlorophenol
2,6-Dichlorophenol
a,a-Dimethyl-
  phenethylamine
2,4-Dimethylphenol
Hexachlorobutadiene
Isophorone
2-Methylnaphtha!ene
Naphthalene
Nitrobenzene
Nitrobenzene-d8 (surr.)
2-Nitrophenol
N-Nitrosodibutyl amine
N-Nitrosopiperidine
1,2,4-Trichlorobenzene
Acenaphthene
Acenaphthylene
1-Chloronaphthalene
2-Chloronaphthalene
4-Chlorophenyl
  phenyl ether
Dibenzofuran
Diethyl phthalate
Dimethyl phthalate
2,4-Dinitrophenol
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Fluorene
2-Fluorobiphenyl
  (surr.)
Hexachlorocyclo-
  pentadiene
1-Naphthylamine
2-Naphthylamine
2-Nitroaniline
3-Nitroaniline
4-Nitroaniline
4-Nitrophenol
Pentachlorobenzene
1,2,4,5-Tetra-
  chlorobenzene
2,3,4,6-Tetra-
  chlorophenol
2,4,6-Tribromo-
  phenol  (surr.)
2,4,6-Trichloro-
  phenol
2,4,5-Trichloro-
  phenol
 (surr.)  =  surrogate
                                   8270A - 35
                                       Revision  1
                                        July  1992

-------
                                   TABLE 5.
                                  (Continued)
Phenanthrene-d
              10
Chrysene-d12
Perylene-d
                                        12
4-Aminobiphenyl
Anthracene
4-Bromophenyl phenyl ether
Di-n-butyl phthalate
4,6-Dinitro-2-methylphenol
Diphenylamine
1,2-Diphenylhydrazine
Fluoranthene
Hexachlorobenzene
N-Nitrosodiphenylamine
Pentachlorophenol
Pentachloroni trobenzene
Phenacetin
Phenanthrene
Pronamide
Benzidine
Benzo(a)anthracene
Bis(2-ethylhexyl) phthalate
Butyl benzyl phthalate
Chrysene
3,3'-Dichlorobenzidine
p-Dimethyl aminoazobenzene
Pyrene
Terphenyl-d14 (surr.)
Benzo(b)fluor-
  anthene
Benzo(k)fluor-
  anthene
Benzo(g,h,i)
  perylene
Benzo(a)pyrene
Dibenz(a,j)acridine
Dibenz(a,h)
  anthracene
7,12-Dimethylbenz-
  (a)anthracene
Di-n-octyl phthalate
Indeno(l,2,3-cd)
  pyrene
3-Methylchol-
  anthrene
(surr.) = surrogate
                                   8270A -  36
                                      Revision  1
                                       July  1992

-------
       TABLE 6.
QC ACCEPTANCE CRITERIA8
Compound
Acenaphthene
Acenaphthylene
Aldrin
Anthracene
Benz(a)anthracene
Benzo(b)fluoranthene
Benzo(k)fluoranthene
Benzo(a)pyrene
Benzo(ghi )perylene
Benzyl butyl phthalate
/3-BHC
6-BHC
Bis(2-chloroethyl) ether
Bis(2-chloroethoxy)methane
Bis(2-chloroisopropyl)ether
Bis(2-ethylhexyl) phthalate
4-Bromophenyl phenyl ether
2-Chloronaphthalene
4-Chlorophenyl phenyl ether
Chrysene
4,4'-DDD
4,4'-DDE
4,4'-DDT
Dibenzo( a, h) anthracene
Di-n-butyl phthalate
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
3,3'-Dichlorobenzidine
Dieldrin
Diethyl phthalate
Dimethyl phthalate
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Di-n-octyl phthalate
Endosulfan sulfate
Endrin aldehyde
Fluoranthene
Fluorene
Heptachlor
Heptachlor epoxide
Hexachlorobenzene
Hexachl orobutad i ene
Test
cone.
(M9/L)
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
Limit
for s
(M9/L)
27.6
40.2
39.0
32.0
27.6
38.8
32.3
39.0
58.9
23.4
31.5
21.6
55.0
34.5
46.3
41.1
23.0
13.0
33.4
48.3
31.0
32.0
61.6
70.0
16.7
30.9
41.7
32.1
71.4
30.7
26.5
23.2
21.8
29.6
31.4
16.7
32.5
32.8
20.7
37.2
54.7
24.9
26.3
Range
for x
(M9/L)
60.1-132.3
53.5-126.0
7.2-152.2
43.4-118.0
41.8-133.0
42.0-140.4
25.2-145.7
31.7-148.0
D-195.0
D-139.9
41.5-130.6
D-100.0
42.9-126.0
49.2-164.7
62.8-138.6
28.9-136.8
64.9-114.4
64.5-113.5
38.4-144.7
44.1-139.9
D-134.5
19.2-119.7
D-170.6
D-199.7
8.4-111.0
48.6-112.0
16.7-153.9
37.3-105.7
8.2-212.5
44.3-119.3
D-100.0
D-100.0
47.5-126.9
68.1-136.7
18.6-131.8
D-103.5
D-188.8
42.9-121.3
71.6-108.4
D-172.2
70.9-109.4
7.8-141.5
37.8-102.2
Range
P» Ps
(%)
47-145
33-145
D-166
27-133
33-143
24-159
11-162
17-163
D-219
D-152
24-149
D-110
12-158
33-184
36-166
8-158
53-127
.60-118
25-158
17-168
D-145
4-136
D-203
D-227
1-118
32-129
D-172
20-124
D-262
29-136
D-114
D-112
39-139
50-158
4-146
D-107
D-209
26-137
59-121
D-192
26.155
D-152
24-116
       8270A -  37
Revision 1
 July 1992

-------
                                   TABLE 6.
                                  (Continued)


Compound
Test Limit
cone. for s
(/ig/L) (/ig/L)
Range
for x
(M9/L)
Range
P» Ps
(%)
Hexachloroethane             100      24.5
Indeno(l,2,3-cd)pyrene       100      44.6
Isophorone                   100      63.3
Naphthalene                  100      30.1
Nitrobenzene                 100      39.3
N-Nitrosodi-n-propylamine    100      55.4
PCB-1260                     100      54.2
Phenanthrene                 100      20.6
Pyrene                       100      25.2
1,2,4-Trichlorobenzene       100      28.1
4-Chloro-3-methylphenol       100      37.2
2-Chlorophenol               100      28.7
2,4-Chlorophenol             100      26.4
2,4-Dimethylphenol           100      26.1
2,4-Dinitrophenol            100      49.8
2-Methyl-4,6-dinitrophenol   100      93.2
2-Nitrophenol                100      35.2
4-Nitrophenol                100      47.2
Pentachlorophenol            100      48.9
Phenol                       100      22.6
2,4,6-Trichlorophenol        100      31.7
55.2-100.0
D-150.9
46.6-180.2
35.6-119.6
54.3-157.6
13.6-197.9
19.3-121.0
65.2-108.7
69.6-100.0
57.3-129.2
40.8-127.9
36.2-120.4
52.5-121.7
41.8-109.0
D-172.9
53.0-100.0
45.0-166.7
13.0-106.5
38.1-151.8
16.6-100.0
52.4-129.2
40-113
D-171
21-196
21-133
35-180
D-230
D-164
54-120
52-115
44-142
22-147
23-134
39-135
32-119
D-191
D-181
29-182
D-132
14-176
5-112
37-144
s

X

p.

D

a
      Standard deviation of four  recovery  measurements,  in  M9/L.

      Average recovery for four recovery measurements,  in M9A-

      Percent recovery measured.

      Detected; result must be greater than  zero.

Criteria from 40 CFR  Part  136  for Method  625.   These  criteria are based
directly on the method performance data in Table 7.  Where necessary, the
limits for  recovery  have been broadened  to assure  applicability  of the
limits to concentrations below those used  to develop Table  7.
                                  8270A - 38
                                                              Revision 1
                                                               July 1992

-------
                         TABLE 7.
METHOD ACCURACY AND PRECISION AS FUNCTIONS OF CONCENTRATION"
Compound
Acenaphthene
Acenaphthylene
Aldrin
Anthracene
Benz(a)anthracene
Chloroethane
Benzo(b)fluoranthene
Benzo ( k) f 1 uoranthene
Benzo(a)pyrene
Benzo(ghi)perylene
Benzyl butyl phthalate
j9-BHC
6-BHC
Bis(2-chloroethyl) ether
Bis(2-chloroethoxy)methane
Bis(2-chloroisopropyl) ether
Bis(2-ethylhexyl) phthalate
4-Bromophenyl phenyl ether
2-Chloronaphthalene
4-Chlorophenyl phenyl ether
Chrysene
4,4'-DDD
4,4'-DDE
4,4'-DDT
Dibenzo (a, h) anthracene
Di-n-butyl phthalate
1,2-Dichlorobenzene
1 ,3-Dichlorobenzene
1,4-Dichlorobenzene
3,3'-Dichlorobenzidine
Dieldrin
Diethyl phthalate
Dimethyl phthalate
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Di-n-octyl phthalate
Endosulfan sulfate
Endrin aldehyde
Fl uoranthene
Fluorene
Heptachlor
Heptachlor epoxide


Accuracy, as
recovery, x'
(M9/L)
0.96C+0.19
0.89C+0.74
0.78C+1.66
0.80C+0.68
0.88C-0.60
0.99C-1.53
0.93C-1.80
0.87C-1.56
0.90C-0.13
0.98C-0.86
0.66C-1.68
0.87C-0.94
0.29C-1.09
0.86C-1.54
1.12C-5.04
1.03C-2.31
0.84C-1.18
0.91C-1.34
0.89C+0.01
0.91C+0.53
0.93C-1.00
0.56C-0.40
0.70C-0.54
0.79C-3.28
0.88C+4.72
0.59C+0.71
0.80C+0.28
0.86C-0.70
0.73C-1.47
1.23C-12.65
0.82C-0.16
0.43C+1.00
0.20C+1.03
0.92C-4.81
1.06C-3.60
0.76C-0.79
0.39C+0.41
0.76C-3.86
0.81C+1.10
0.90C-0.00
0.87C-2.97
0.92C-1.87
8270A - 39

Single analyst
precision, s'
(M9/L)
O.lBx-0.12
0.24X-1.06
0.27X-1.28
0.21X-0.32
O.lSx+0.93
0.14X-0.13
0.22X+0.43
0.19X+1.03
0.22X+0.48
0.29X+2.40
0.18X+0.94
0.20X-0.58
0.34X+0.86
0.35X-0.99
0.16X+1.34
0.24X+0.28
0.26X+0.73
O.lSx+0.66
0.07X+0.52
0.20X-0.94
0.28X+0.13
0.29X-0.32
0.26X-1.17
0.42X+0.19
0.30X+8.51
0.13X+1.16
0.20X+0.47
0.25X+0.68
0.24X+0.23
0.28X+7.33
0.20X-0.16
0.28X+1.44
0.54X+0.19
0.12x+1.06
0.14X+1.26
0.21X+1.19
0.12X+2.47
O.lSx+3.91
0.22X-0.73
0.12X+0.26
0.24X-0.56
0.33X-0.46


Overall
precision,
S' (M9/L)
0.21X-0.67
0.26x-0.54
0.43X+1.13
0.27X-0.64
0.26X-0.21
0.17X-0.28
0.29X+0.96
0.35X+0.40
0.32X+1.35
O.Slx-0.44
0.53X+0.92
0.30X+1.94
0.93X-0.17
0.35X+0.10
0.26X+2.01
0.25X+1.04
0.36X+0.67
0.16X+0.66
0.13X+0.34
0.30X-0.46
0.33X-0.09
0.66X-0.96
0.39X-1.04
0.65X-0.58
0.59X+0.25
0.39X+0.60
0.24X+0.39
0.41X+0.11
0.29X-I-0.36
0.47X+3.45
0.26X-0.07
0.52X+0.22
1.05X-0.92
0.21X+1.50
0.19X+0.35
0.37X+1.19
0.63X-1.03
0.73X-0.62
0.28X-0.60
0.13X+0.61
0.50X-0.23
0.28X+0.64
Revision 1
July 1992

-------




Compound
Hexachlorobenzene
Hexachlorobutadiene
Hexachloroethane
Indeno(l,2,3-cd)pyrene
Isophorone
Naphthalene
Nitrobenzene
N-Nitrosodi-n-propylamine
PCB-1260
Phenanthrene
Pyrene
1,2,4-Trichlorobenzene
4-Chl oro-3-methyl phenol
2-Chlorophenol
2,4-Dichlorophenol
2, 4-Dimethyl phenol
2,4-Dinitrophenol
2-Methyl-4,6-dinitrophenol
2-Nitrophenol
4-Nitrophenol
Pentachlorophenol
Phenol
2,4,6-Trichlorophenol
TABLE 7.
(Continued)
Accuracy, as
recovery, x'
(M9/L)
0.74C+0.66
0.71C-1.01
0.73C-0.83
0.78C-3.10
1.12C+1.41
0.76C+1.58
1.09C-3.05
1.12C-6.22
0.81C-10.86
0.87C+0.06
0.84C-0.16
0.94C-0.79
0.84C+0.35
0.78C+0.29
0.87C-0.13
0.71C+4.41
0.81C-18.04
1.04C-28.04
0.07C-1.15
0.61C-1.22
0.93C+1.99
0.43C+1.26
0.91C-0.18


Single analyst
precision, s '
(M9/L)
O.lSx-0.10
0.19X+0.92
O.Ux+0.67
0.29X+1.46
0.27x+0.77
0.21X-0.41
0.19X+0.92
0.27X+0.68
0.35X+3.61
0.12X+0.57
0.16X+0.06
O.lSx+0.85
0.23X+0.75
O.lSx+1.46
O.lBx+1.25
Q.16X+1.21
0.38X+2.36
O.lOx+42.29
0.16X+1.94
0.38X+2.57
0.24X+3.03
0.26X+0.73
O.lSx+2.22


Overall
precision,
S' (M9/L)
0.43X-0.52
0.26X+0.49
0.17x+0.80
O.BOx-0.44
0.33x+0.26
0.30X-0.68
0.27X+0.21
0.44X+0.47
0.43X+1.82
O.lSx+0.25
O.lSx+0.31
0.21X+0.39
0.29X+1.31
0.28X+0.97
0.21X-I-1.28
0.22X+1.31
0.42X+26.29
0.26X+23.10
0.27X+2.60
0.44X+3.24
0.30x^4.33
0.35X+0.58
0.22X+1.81
X'
S'


C

x"
Expected  recovery  for  one  or  more  measurements  of  a  sample
containing a concentration of C, in M9/L.

Expected single  analyst  s_tandard deviation of  measurements  at an
average concentration of x, in M9/L.

Expected interlaboratory  standajd  deviation of  measurements  at an
average concentration found of x, in /j.g/1.

True value for the concentration, in M9/L.

Average  recovery found for  measurements of  samples  containing a
concentration of C, in M9/L-
                                  8270A - 40
                                                        Revision 1
                                                         July 1992

-------
                                   TABLE 8.
      SURROGATE  SPIKE  RECOVERY  LIMITS  FOR WATER AND  SOIL/SEDIMENT SAMPLES
                              Low/High                  Low/High
Surrogate Compound              Water                 Soil/Sediment
Nitrobenzene-d5                 35-114                   23-120
2-Fluorobiphenyl               43-116                   30-115
p-Terphenyl-du                 33-141                   18-137

Phenol-d,                       10-94                    24-113
2-Fluorophenol                  21-100                   25-121
2,4,6-Tribromophenol           10-123                   19-122
                                  8270A - 41                        Revision 1
                                                                     July 1992

-------
                   fcIC
                        /-86  6:26:88
                   S«*Pl£: BASE (CIO STO,2VJL/2uHC UL
                   CONDS.t
                   RHNCCi C   1,2780  LflttLj N  6,  4.0  UUnN:
CUTrt:  5lbH£ttH»786 kl
CM.lt  51BHS666786 13
         0, I. d J  6
                                                                                  2t)6 TO 2766
                           U 2d-  3
                          CD
                          3>
                          o
                          DC
                          o
  oo
  ro
             RIC
 i ft)
  <
                           ~T	
                            566
                            8:26
                                                                                                                                   oo
                                                                                                                                   O

                                                                                                                                   O
                                                                                                                                   oo
                                                                                                    AJL
                                                                              o
                                                                              3>
                                                                              X)
                                                                              O
                                —I	
                                 2W9
                                 33:26
25*
41:46
                                                                                                                     in;
VO
10
ro

-------
                                         METHOD 8270A
                            SEMIVOLATILE ORGANIC COMPOUNDS  BY
    GAS CHROMATOGRAPHY/MASS  SPECTROMETRY (GC/MS): CAPILLARY  COLUMN TECHNIQUE
 ^  1 Prepare
*ampl« uiing
 Method 3540
  or 3550
 7  1 Prepare
sample uiing
 Method 3S10
  or 3520
                        7 1  Prepare
                       samp 1• using
                       Method 3540.
                       3550  or 3580
  1 S 1 Screen
extract on CC/fID
  or CC/PID to
eliminate sampl«»
  that are too
  concent ra ted
                        7 2  Cleanup
                         ext ract
                                       753  Analyze
                                     extract  by GC/MS.
                                     using appripriate
                                       fused•aiitea
                                     capi1i a ry column
                754 Dilute
                   • M tract
                                                                                          So
                       7 3 Set CC/MS
                         opera ting
                        condit L onj
                      Perform  initia1
                       calibration
                                       ? 6 1  'r^ufy
                                        a r. a . \ '. « o y
                                       ca^urf ; i -9 *,Ke
                                     sample ana  standard
                                       .•nass  9 o e c '. r a
                      7 4 Perform daily
                      calibration with
                      SPCCs »nd CCC»
                      prior to ana lysis
                        of sample*
                                       7 6 2 Caic-iat*
                                      concentration of
                                       eac!*. individual
                                       analyle  3 e u o r t
                                         results
                                                                                        Stop
                                            8270A - 43
                                          Revision 1
                                           July  IS
                                                                                                 32

-------
06
O
W

-------
                                 METHOD 8270B

                       SEMIVOLATILE ORGANIC  COMPOUNDS BY
   GAS CHROMATOGRAPHY/MASS SPECTROMETRV (GC/MS): CAPILLARY COLUMN TECHNIQUE
1.0   SCOPE AND APPLICATION

      1.1   Method 8270 is used to determine the concentration of semivolatile
organic compounds in extracts prepared from all  types of solid waste matrices,
soils, and ground water.  Direct  injection  of a  sample  may  be used in limited
applications.  The following compounds can be determined by this method:
Compounds
        Appropriate Preparation Techniques


CAS Noa      3510
      3540/
3520  3541  3550  3580
Acenaphthene
Acenaphthene-d10 (I.S.)
Acenaphthylene
Acetophenone
2-Acetylaminofluorene
l-Acetyl-2-thiourea
Aldrin
2-Aminoanthraquinone
Aminoazobenzene
4-Aminobiphenyl
3-Amino-9-ethylcarbazole
Anilazine
Aniline
o-Anisidine
Anthracene
Aramite
Aroclor - 1016
Aroclor - 1221
Aroclor - 1232
Aroclor - 1242
Aroclor - 1248
Aroclor - 1254
Aroclor - 1260
Azinphos-methyl
Barban
Benzidine
Benzoic acid
Benz(a)anthracene
Benzo(b)fluoranthene
Benzo(k)fluoranthene
Benzo (g , h , i ) peryl ene
Benzo(a)pyrene
83-32-9

208-96-8
98-86-2
53-96-3
591-08-2
309-00-2
117-79-3
60-09-3
92-67-1
132-32-1
101-05-3
62-53-3
90-04-0
120-12-7
140-57-8
12674-11-2
11104-28-2
11141-16-5
53469-21-9
12672-29-6
11097-69-1
11096-82-5
86-50-0
101-27-9
92-87-5
65-85-0
56-55-3
205-99-2
207-08-9
191-24-2
50-32-8
X
X
X
X
X
LR
X
X
X
X
X
X
X
X
X
HS(43)
X
X
X
X
X
X
X
HS(62)
LR
CP
X
X
X
X
X
X
X
X
X
ND
NO
ND
X
ND
ND
ND
X
ND
X
ND
X
ND
X
X
X
X
X
X
X
ND
ND
CP
X
X
X
X
X
X
X
X
X
ND
ND
ND
X
ND
ND
ND
ND
ND
ND
ND
X
ND
X
X
X
X
X
X
X
ND
ND
CP
ND
X
X
X
X
X
X
X
X
ND
ND
ND
X
ND
ND
ND
ND
ND
X
ND
X
ND
X
X
X
X
X
X
X
ND
ND
CP
X
X
X
X
X
X
X
X
X
X
X
LR
X
X
X
X
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
LR
CP
X
X
X
X
X
X
                                  8270B  - 1
                                   Revision  2
                               September 1994

-------
Appropriate Preparation Techniques
Compounds
p-Benzoquinone
Benzyl alcohol
a-BHC
jS-BHC
5-BHC
-y-BHC (Lindane)
Bis(2-chloroethoxy)methane
Bis(2-chloroethyl) ether
Bis(Z-chloroisopropyl) ether
Bis(2-ethylhexyl) phthalate
4-Bromophenyl phenyl ether
Bromoxynil
Butyl benzyl phthalate
2-sec-Butyl-4,6-dinitrophenol
Captafol
Captan
Carbaryl
Carbofuran
Carbophenothion
Chlordane
Chlorfenvinphos
4-Chloroaniline
Chlorobenzilate
5-Chloro- 2 -methyl aniline
4- Chi oro-3 -methyl phenol
3-(Chloromethyl )pyridine
hydrochloride
1-Chloronaphthalene
2-Chloronaphthalene
2-Chlorophenol
4-Chloro- 1,2- phenyl enedi ami ne
4 -Chi oro- 1,3 -phenyl enedi ami ne
4-Chlorophenyl phenyl ether
Chrysene
Chrysene-d12 (I.S.)
Coumaphos
p-Cresidine
Crotoxyphos
2-Cyclohexyl-4,6-dinitro-phenol
4,4'-DDD
4,4'-DDE
4,4'-DDT
Demeton-0
Demeton-S
Oiallate (cis or trans)
2,4-Diaminotoluene
CAS Noa
106-51-4
100-51-6
319-84-6
319-85-7
319-86-8
58-89-9
111-91-1
111-44-4
108-60-1
117-81-7
101-55-3
1689-84-5
85-68-7
88-85-7
2425-06-1
133-06-2
63-25-2
1563-66-2
786-19-6
57-74-9
470-90-6
106-47-8
510-15-6
95-79-4
59-50-7

6959-48-4
90-13-1
91-58-7
95-57-8
95-83-0
5131-60-2
7005-72-3
218-01-9

56-72-4
120-71-8
7700-17-6
131-89-5
72-54-8
72-55-9
50-29-3
298-03-3
126-75-0
2303-16-4
95-80-7
3510
OE
X
X
X
X
X
X
X
X
X
X
X
X
X
HS(55)
HS(40)
X
X
X
X
X
X
X
X
X

X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
HS(68)
X
X
DC,OE(42)
3520
ND
X
X
X
X
X
X
X
X
X
X
ND
X
ND
ND
ND
ND
ND
ND
X
ND
ND
ND
ND
X

ND
X
X
X
X
X
X
X
X
ND
ND
ND
ND
X
X
X
ND
ND
ND
ND
3540/
3541
ND
ND
X
X
X
X
X
X
X
X
X
ND
X
ND
ND
ND
ND
ND
ND
X
ND
ND
ND
ND
X

ND
X
X
X
ND
ND
X
X
X
ND
ND
ND
ND
X
X
X
ND
ND
ND
ND
3550
ND
X
X
X
X
X
X
X
X
X
X
ND
X
ND
ND
ND
ND
ND
ND
X
ND
ND
ND
ND
X

ND
X
X
X
ND
ND
X
X
X
ND
ND
ND
ND
X
X
X
ND
ND
ND
ND
3580
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X

X
X
X
X
ND
ND
X
X
X
X
X
X
LR
X
X
X
X
X
X
X
8270B - 2
    Revision 2
September 1994

-------
Compounds
        Appropriate Preparation Techniques


CAS No8      3510
      3540/
3520  3541 3550   3580
Dibenz(a,j)acridine
Dibenz( a, h) anthracene
Dibenzofuran
Dibenzo(a,e)pyrene
l,2-Dibromo-3-chloropropane
Di-n-butyl phthalate
Dichlone
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
l,4-Dichlorobenzene-d4 (I.S)
3,3'-Dich1orobenzidine
2,4-Dichlorophenol
2,6-Dichlorophenol
Dichlorovos
Dicrotophos
Dieldrin
Diethyl phthalate
Diethylstilbestrol
Diethyl sulfate
Dihydrosaffrole
Dimethoate
3,3' -Dimethoxybenzidine
Dimethyl aminoazobenzene
7,12-Dimethylbenz(a)-
anthracene
3,3'-Dimethylbenzidine
a,a-Dimethylphenethylamine
2,4-Dimethylphenol
Dimethyl phthalate
1,2-Dinitrobenzene
1,3-Dinitrobenzene
1,4-Dinitrobenzene
4, 6-Dinitro-2-methyl phenol
2,4-Dinitrophenol
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Dinocap
Dinoseb
Dioxathion
Diphenylamine
5,5-Diphenylhydantoin
1,2-Diphenylhydrazine
Di-n-octyl phthalate
Disulfoton
224-42-0
53-70-3
132-64-9
192-65-4
96-12-8
84-74-2
117-80-6
95-50-1
541-73-1
106-46-7

91-94-1
120-83-2
87-65-0
62-73-7
141-66-2
60-57-1
84-66-2
56-53-1
64-67-5
56312-13-1
60-51-5
119-90-4
60-11-7
57-97-6

119-93-7
122-09-8
105-67-9
131-11-3
528-29-0
99-65-0
100-25-4
534-52-1
51-28-5
121-14-2
606-20-2
39300-45-3
88-85-7
78-34-2
122-39-4
57-41-0
122-66-7
117-84-0
298-04-4
X
X
X
ND
X
X
OE
X
X
X
X
X
X
X
X
X
X
X
AW,OS(67)
LR
ND
HE,HS(31)
X
X
CP(45)

X
ND
X
X
X
X
HE(14)
X
X
X
X
CP,HS(28)
X
ND
X
X
X
X
X
ND
X
X
ND
X
X
ND
X
X
X
X
X
X
ND
ND
ND
X
X
ND
ND
ND
ND
ND
ND
ND

ND
ND
X
X
ND
ND
ND
X
X
X
X
ND
ND
ND
X
ND
X
X
ND
ND
X
ND
ND
ND
X
ND
X
X
X
X
X
X
ND
ND
ND
X
X
ND
ND
ND
ND
ND
ND
ND

ND
ND
X
X
ND
ND
ND
X
X
X
X
ND
ND
ND
X
ND
X
X
ND
ND
X
X
ND
ND
X
ND
X
X
X
X
X
X
ND
ND
ND
X
X
ND
ND
ND
ND
ND
ND
ND

ND
ND
X
X
ND
ND
ND
X
X
X
X
ND
ND
ND
X
ND
X
X
ND
X
X
X
X
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
X
LR
ND
X
LR
X
CP

X
X
X
X
X
X
X
X
X
X
X
CP
X
ND
X
X
X
X
X
                                  8270B - 3
                                   Revision  2
                               September  1994

-------
Compounds
        Appropriate Preparation_Technicities


CAS No8      3510
      3540/
3520  3541 3550   3580
Endosulfan I
Endosulfan II
Endosulfan sulfate
Endrin
Endrin aldehyde
Endrin ketone
EPN
Ethion
Ethyl carbamate
Ethyl methanesulfonate
Ethyl parathion
Famphur
Fensulfothion
Fenthion
Fluchloral in
Fluoranthene
Fluorene
2-Fluorobiphenyl (surr.)
2-Fluorophenol (surr.)
Heptachlor
Heptachlor epoxide
Hexachlorobenzene
Hexachl orobutad i ene
Hexachlorocyclopentadiene
Hexachl oroethane
Hexachl orophene
Hexachl oropropene
Hexamethyl phosphoramide
Hydroquinone
Indeno(l,2,3-cd)pyrene
Isodrin
Isophorone
Isosafrole
Kepone
Leptophos
Malathion
Maleic anhydride
Mestranol
Methapyrilene
Methoxychlor
3-Methyl chol anthrene
4,4'-Methylenebis
(2-chloroaniline)
4,4'-Methylenebis
(N,N-dimethylaniline)
959-98-8
33213-65-9
1031-07-8
72-20-8
7421-93-4
53494-70-5
2104-64-5
563-12-2
51-79-6
62-50-0
56-38-2
52-85-7
115-90-2
55-38-9
33245-39-5
206-44-0
86-73-7
321-60-8
367-12-4
76-44-8
1024-57-3
118-74-1
87-68-3
77-47-4
67-72-1
70-30-4
1888-71-7
680-31-9
123-31-9
193-39-5
465-73-6
78-59-1
120-58-1
143-50-0
21609-90-5
121-75-5
108-31-6
72-33-3
91-80-5
72-43-5
56-49-5

101-14-4

101-61-1
X
X
X
X
X
X
X
X
DC(28)
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
AW,CP(62)
X
X
ND
X
X
X
DC(46)
X
X
HS(5)
HE
X
X
X
X

OE,OS(0)

X
X
X
X
X
X
X
ND
ND
ND
ND
X
ND
ND
ND
ND
X
X
X
X
X
X
X
X
X
X
ND
ND
ND
ND
X
ND
X
ND
ND
ND
ND
ND
ND
ND
ND
ND

ND

X
X
X
X
X
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
X
X
X
X
X
X
X
X
X
X
ND
ND
ND
ND
X
ND
X
ND
ND
ND
ND
ND
ND
ND
ND
ND

ND

ND
X
X
X
X
X
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
X
X
X
X
X
X
X
X
X
X
ND
ND
ND
ND
X
ND
X
ND
ND
ND
ND
ND
ND
ND
ND
ND

ND

ND
X
X
X
X
X
X
X
X
X
X
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
X
CP
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X

LR

ND
                                   8270B - 4
                                   Revision  2
                               September  1994

-------
Compounds
        Appropriate Preparation Techniques


CAS Noa      3510
      3540/
3520  3541 3550   3580
Methyl methanesulfonate
2-Methyl naphthalene
2-Methyl-5-nitroanil ine
Methyl parathion
2-Methylphenol
3-Methyl phenol
4-Methyl phenol
2-Methylpyridine
Mevinphos
Mexacarbate
Mi rex
Monocrotophos
Naled
Naphthalene
Naphthalene-d8 (I.S.)
1,4-Naphthoquinone
1-Naphthylamine
2-Naphthylamine
Nicotine
5-Nitroacenaphthene
2-Nitroanil ine
3-Nitroanil ine
4-Nitroanil ine
5-Nitro-o-anisidine
Nitrobenzene
Nitrobenzene-d5 (surr.)
4-Nitrobiphenyl
Nitrofen
2-Nitrophenol
4-Nitrophenol
5-Nitro-o-toluidine
Nitroquinoline-1 -oxide
N -N itrosodi butyl ami ne
N-Nitrosodi ethyl ami ne
N-Nitrosodimethylamine
N-Nitrosomethyl ethyl ami ne
N-Nitrosodiphenylamine
N-Nitrosodi -n-propylamine
N-Nitrosomorphol ine
N-Nitrosopiperidine
N-Nitrosopyrrol idine
Octamethyl pyrophosphoramide
4,4'-Oxydianil ine
Parathion
Pentachl orobenzene
66-27-3
91-57-6
99-55-8
298-00-0
95-48-7
108-39-4
106-44-5
109-06-8
7786-34-7
315-18-4
2385-85-5
6923-22-4
300-76-5
91-20-3

130-15-4
134-32-7
91-59-8
54-11-5
602-87-9
88-74-4
99-09-2
100-01-6
99-59-2
98-95-3

92-93-3
1836-75-5
88-75-5
100-02-7
99-55-8
56-57-5
924-16-3
55-18-5
62-75-9
10595-95-6
86-30-6
621-64-7
59-89-2
100-75-4
930-55-2
152-16-9
101-80-4
56-38-2
608-93-5
X
X
X
X
X
X
X
X
X
HE,HS(68)
X
HE
X
X
X
X
OS(44)
X
DE(67)
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
ND
X
X
LR
X
X
X
ND
X
X
ND
ND
ND
ND
X
ND
ND
ND
ND
ND
X
X
ND
ND
ND
ND
ND
X
X
X
ND
X
X
ND
ND
X
X
ND
ND
ND
ND
X
ND
X
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
X
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
X
X
ND
ND
X
X
ND
ND
ND
ND
X
ND
X
X
ND
ND
ND
ND
ND
ND
ND
ND
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
X
X
ND
ND
ND
ND
ND
X
X
X
ND
X
X
ND
ND
X
X
ND
ND
ND
ND
X
ND
X
X
ND
ND
ND
ND
ND
ND
ND
X
X
ND
X
X
X
X
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
LR
X
X
X
                                  8270B  - 5
                                   Revision 2
                               September 1994

-------
Compounds
        Appropriate Preparation Techniques


CAS No"      3510
      3540/
3520  3541 3550   3580
Pentach.1 oroni trobenzene
Pentachlorophenol
Perylene-d12 (I.S.)
Phenacetin
Phenanthrene
Phenanthrene-d10 (I.S.)
Phenobarbital
Phenol
Phenol -d6 (surr.)
1,4-Phenylenediamine
Phorate
Phosalone
Phosmet
Phosphamidon
Phthalic anhydride
2-Picol ine
Piperonyl sulfoxide
Pronamide
Propylthiouracil
Pyrene
Pyridine
Resorcinol
Safrole
Strychnine
Sul fall ate
Terbufos
Terphenyl-d14(surr.)
1,2,4 , 5-Tetrachl orobenzene
2,3,4,6-Tetrachlorophenol
Tetrachlorvinphos
Tetraethyl dithiopyrophosphate
Tetraethyl pyrophosphate
Thionazine
Thiophenol (Benzenethiol )
Toluene diisocyanate
o-Toluidine
Toxaphene
2,4,6-Tribromophenol (surr.)
1,2, 4 -Trichl orobenzene
2,4,5-Trichlorophenol
2,4,6-Trichlorophenol
Trifluralin
2,4,5-Trimethylaniline
Trimethyl phosphate
82-68-8
87-86-5

62-44-2
85-01-8

50-06-6
108-95-2

106-50-3
298-02-2
2310-17-0
732-11-6
13171-21-6
85-44-9
109-06-8
120-62-7
23950-58-5
51-52-5
129-00-0
110-86-1
108-46-3
94-59-7
60-41-3
95-06-7
13071-79-9
1718-51-0
95-94-3
58-90-2
961-11-5
3689-24-5
107-49-3
297-97-2
108-98-5
584-84-9
95-53-4
8001-35-2

120-82-1
95-95-4
88-06-2
1582-09-8
137-17-7
512-56-1
X
X
X
X
X
X
X
DC(28)
DC(28)
X
X
HS(65)
HS(15)
HE(63)
CP,HE(1)
ND
X
X
LR
X
ND
DC,OE(10)
X
AW,OS(55)
X
X
X
X
X
X
X
X
X
X
HE(6)
X
X
X
X
X
X
X
X
HE(60)
ND
X
X
ND
X
X
ND
X
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
X
ND
ND
ND
ND
ND
ND
X
ND
ND
ND
X
ND
ND
ND
ND
ND
X
X
X
X
X
ND
ND
ND
ND
X
X
ND
X
X
ND
X
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
X
X
X
ND
X
ND
ND
ND
ND
X
X
ND
X
X
ND
X
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
X
ND
ND
ND
ND
ND
ND
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
X
X
X
X
X
ND
ND
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
X
CP
ND
X
X
LR
X
ND
X
X
X
X
X
X
X
X
X
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
                                   8270B  -  6
                                   Revision 2
                               September 1994

-------
                                         Appropriate Preparation Techniques

                                                             3540/
Compounds                        CAS No8      3510     3520   3541 3550   3580
1,3,5-Trinitrobenzene 99-35-4
Tris(2,3-dibromopropyl) phosphate 126-72-7
Tri-p-tolyl phosphate 78-32-0
0,0,0-Triethyl phosphorothioate 126-68-1
X
X
X
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
X
LR
X
X
a     Chemical Abstract Service Registry Number.

AW =  Adsorption to walls of glassware during extraction and storage.
CP =  Nonreproducible chromatographic performance.
DC =  Unfavorable  distribution  coefficient (number  in  parenthesis  is percent
      recovery).
HE =  Hydrolysis during  extraction accelerated by  acidic  or  basic  conditions
      (number in parenthesis is percent recovery).
HS =  Hydrolysis during storage (number in parenthesis  is percent stability).
LR =  Low response.
ND =  Not determined.
OE =  Oxidation during  extraction accelerated by basic  conditions  (number in
      parenthesis  is percent recovery).
OS =  Oxidation during storage  (number in parenthesis is percent stability).
X =   Greater than 70 percent recovery by this technique.


      1.2   Method  8270  can be used  to  quantitate most neutral,  acidic,  and
basic organic compounds that are  soluble  in methylene  chloride and  capable of
being eluted without derivatization as sharp  peaks  from a  gas  chromatographic
fused-silica  capillary  column  coated  with  a slightly  polar   silicone.   Such
compounds include polynuclear aromatic hydrocarbons, chlorinated  hydrocarbons and
pesticides,  phthalate esters, organophosphate  esters, nitrosamines, haloethers,
aldehydes,  ethers,  ketones, anilines,  pyridines,  quinolines,   aromatic  nitro
compounds,  and  phenols,  including nitrophenols.   See  Table  1 for a  list of
compounds and  their characteristic ions that have been evaluated  on the specified
GC/MS system.

      1.3   The following  compounds  may require special treatment  when  being
determined by  this  method.   Benzidine can be subject to oxidative losses during
solvent  concentration.   Also,  chromatography  is  poor.   Under  the  alkaline
conditions of  the extraction step,  a-BHC, 7-BHC, Endosulfan  I and II, and Endrin
are subject to decomposition.  Neutral  extraction should be performed if these
compounds  are expected.    Hexachlorocyclopentadiene  is  subject  to  thermal
decomposition  in the inlet of the gas chromatograph, chemical reaction in acetone
solution, and  photochemical  decomposition.  N-nitrosodimethylamine is difficult
to separate from the  solvent under the chromatographic conditions  described.
N-nitrosodiphenylaroine decomposes  in the gas chromatographic inlet  and cannot be
separated   from   diphenylamine.       Pentachlorophenol,   2,4-dinitrophenol,


                                   8270B -  7                         Revision 2
                                                                September 1994

-------
4-nitrophenol, 4,6-dinitro-2-methylpheno1,4-chloro-3-methylphenol, benzoicacid,
2-nitroaniline, 3-nitroanil ine, 4-chloroaniline,  and benzyl  alcohol are subject
to erratic chromatographic behavior, especially if the GC system is contaminated
with high boiling material.

      1.4    The   estimated  quantitation  limit   (EQL)   of  Method  8270  for
determining  an  individual  compound is approximately  1  mg/kg  (wet  weight) for
soil/sediment samples, 1-200 mg/kg  for wastes (dependent on matrix and method of
preparation), and 10 /ug/L for ground water samples  (see  Table  2).  EQLs will be
proportionately  higher for  sample extracts  that  require  dilution  to  avoid
saturation of the detector.

      1.5    This  method is restricted to  use  by or  under  the  supervision of
analysts  experienced  in the  use  of gas  chromatograph/mass  spectrometers and
skilled in the interpretation of mass spectra.  Each analyst must demonstrate the
ability to generate acceptable results with this method.


2.0   SUMMARY OF METHOD

      2.1    Prior  to using  this  method, the  samples should be prepared for
chromatography using  the  appropriate sample preparation  and  cleanup  methods.
This  method describes  chromatographic  conditions   that  will  allow  for  the
separation  of  the  compounds  in  the extract  and   for  their qualitative  and
quantitative analysis by mass  spectrometry.


3.0   INTERFERENCES

      3.1    Raw  GC/MS  data  from  all  blanks,   samples,  and  spikes  must  be
evaluated for interferences.  Determine  if the  source of interference is in the
preparation and/or cleanup of the samples  and take corrective action to eliminate
the problem.

      3.2    Contamination  by carryover can occur whenever high-concentration and
low-concentration samples  are  sequentially analyzed.  To reduce carryover, the
sample syringe must  be rinsed out between samples  with  solvent.   Whenever an
unusually  concentrated sample is  encountered,  it  should  be followed by the
analysis of  solvent to check for cross contamination.


4.0   APPARATUS AND MATERIALS

      4.1    Gas chromatograph/mass spectrometer  system

             4.1.1  Gas chromatograph  -  An analytical  system complete with  a
      temperature-programmable   gas   chromatograph   suitable   for   splitless
      injection  and  all required  accessories,  including syringes,  analytical
      columns, and gases.   The capillary column  should  be directly coupled to
      the  source.
                                   8270B -  8                        Revision 2
                                                                September 1994

-------
            4.1.2 Column - 30 m x 0.25 mm ID (or 0.32 mm ID) 1 ^m film thickness
      silicone-coated  fused-silica  capillary column  (J&W Scientific  DB-5  or
      equivalent).

            4.1.3 Mass  spectrometer  -  Capable of scanning from 35  to  500 amu
      every 1  sec or  less,  using 70 volts  (nominal)  electron energy  in the
      electron impact  ionization mode.  The  mass  spectrometer  must  be  capable
      of  producing  a  mass  spectrum  for decaf!uorotriphenylphosphine  (DFTPP)
      which meets all of the criteria in Table 3 when 1 ^L of the GC/MS tuning
      standard is injected through the GC (50 ng of DFTPP).

            4.1.4 GC/MS interface - Any GC-to-MS interface that gives acceptable
      calibration points at  50  ng per  injection  for each compound of interest
      and achieves acceptable tuning performance  criteria  may  be  used.   For a
      narrow-bore capillary  column,  the interface  is usually capillary-direct
      into the mass spectrometer source.

            4.1.5 Data system - A computer system must be interfaced  to the mass
      spectrometer.   The  system must  allow the  continuous acquisition  and
      storage on machine-readable media  of all mass spectra obtained throughout
      the duration  of the  chromatographic  program.   The computer  must  have
      software that can search any GC/MS data file for ions of a specific mass
      and that can plot such  ion  abundances  versus  time  or scan  number.   This
      type of  plot  is  defined as an  Extracted  Ion  Current Profile  (EICP).
      Software must also be available that allows integrating the abundances in
      any EICP between specified  time or scan-number  limits.   The most recent
      version of the EPA/NIST Mass Spectral  Library should also be available.

            4.1.6 Guard column  (optional)  (J&W  Deactivated Fused  Silica,  0.25
      mm ID x 6 m, or equivalent)  between the  injecti'on port and the analytical
      column joined  with  column  joiners  (Hewlett  Packard No.  5062-3556,  or
      equivalent).

      4.2   Syringe  -  10 nl.

      4.3   Volumetric  flasks,  Class A -  Appropriate  sizes  with  ground glass
stoppers.

      4.4   Balance  - Analytical, 0.0001 g.

      4.5   Bottles  - glass with Teflon-lined screw caps or crimp tops.


5.0   REAGENTS

      5.1   Reagent  grade   inorganic  chemicals  shall  be  used  in  all  tests.
Unless otherwise indicated, it is intended that all reagents shall conform to the
specifications of the Committee  on Analytical  Reagents of the American Chemical
Society, where such  specifications are  available.   Other  grades  may be used,
provided it  is  first  ascertained that  the reagent  is  of  sufficiently high purity
to permit its use without lessening  the accuracy of the determination.
                                  8270B  - 9                         Revision 2
                                                                September 1994

-------
      5.2   Organic-free reagent water - All references to water  in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.3   Stock  standard  solutions (1000 mg/L)  -  Standard  solutions can be
prepared from pure standard materials or purchased as certified solutions.

            5.3.1  Prepare stock standard solutions by accurately weighing about
      0.0100 g  of  pure  material.   Dissolve the  material  in pesticide quality
      acetone  or  other  suitable   solvent  and  dilute to  volume in  a  10  ml
      volumetric flask.  Larger volumes  can  be used  at the convenience of the
      analyst.  When compound purity  is assayed to be  96% or greater, the weight
      may be used without correction to calculate the  concentration of the stock
      standard.    Commercially  prepared  stock  standards  may  be used  at  any
      concentration  if  they  are  certified  by  the  manufacturer  or by  an
      independent  source.

            5.3.2  Transfer the stock standard solutions into bottles with Teflon
      lined screw-caps.   Store at -10°C to -20°C or less and protect from light.
      Stock  standard solutions  should  be checked  frequently  for signs  of
      degradation or evaporation, especially just prior to  preparing calibration
      standards  from them.

            5.3.3  Stock  standard  solutions must  be  replaced after 1  year or
      sooner  if comparison  with   quality  control check  samples indicates  a
      problem.

      5.4   Internal standard solutions  - The internal standards recommended are
l,4-dichlorobenzene-d4,   naphthalene-ds,   acenaphthene-d10,   phenanthrene-d10,
chrysene-d12,  and perylene-d12  (see Table 5).   Other  compounds may  be  used as
internal standards  as  long  as the  requirements  given in  Sec. 7.3.2  are met.
Dissolve 0.200  g of each  compound with a  small  volume  of  carbon  disulfide.
Transfer to a 50 ml volumetric flask and dilute to volume with methylene chloride
so that the final  solvent is  approximately  20% carbon disulfide.  Most of the
compounds are also  soluble in  small  volumes  of methanol,  acetone, or  toluene,
except for perylene-d12.   The resulting  solution will contain each standard at
a concentration  of 4,000 ng/>L.  Each  1  mi sample extract undergoing  analysis
should be spiked with 10 /nL  of the internal  standard solution,  resulting in a
concentration of 40 ng/juL of  each  internal  standard.   Store at -10°C  to -20°C
or less when not being used.

      5.5   GC/MS  tuning  standard  - A methylene  chloride solution  containing
50 ng//zL  of decafluorotriphenylphosphine  (DFTPP) should be  prepared.    The
standard should also contain  50 ng/ptL each of 4,4'-DDT, pentachlorophenol,  and
benzidine to verify injection port  inertness and GC column performance.  Store
at -10°C to -20°C or less when not  being used.

      5.6   Calibration  standards  - A  minimum of five calibration  standards
should  be   prepared.    One  of  the  calibration  standards   should  be  at  a
concentration near, but  above,  the method detection  limit;  the  others should
correspond to the range of concentrations found in real samples but  should not
exceed the  working  range of the GC/MS system.  Each standard should contain each
analyte for detection by this method (e.g.  some or all of the compounds listed
in Table 1  may be included).   Each  1 ml  aliquot of calibration standard should


                                  8270B  - 10                        Revision 2
                                                                September 1994

-------
be spiked with 10 juL of the internal standard solution prior to analysis.   All
standards should  be  stored at -10°C to  -20°C  or  less,  and should  be  freshly
prepared once a year, or sooner if check standards indicate a problem.  The daily
calibration standard should be prepared weekly and stored at 4°C.

      5.7    Surrogate  standards   -  The  recommended  surrogate  standards  are
phenol-d6,    2-fluorophenol,    2,4,6-tribromophenol,    nitrobenzene-d5,
2-fluorobiphenyl, and p-terphenyl-du.  See Method 3500 for the instructions on
preparing the surrogate standards. Determine what concentration should be in the
blank extracts after all  extraction, cleanup,  and  concentration steps.   Inject
this concentration into the GC/MS to determine recovery of surrogate standards
in all  blanks, spikes,  and  sample extracts.  Take into account all  dilutions of
sample extracts.

      5.8    Matrix  spike  standards  -  See  Method  3500  for  instructions  on
preparing the matrix spike  standard.  Determine what concentration should be in
the  blank  extracts after  all  extraction, cleanup,  and  concentration  steps.
Inject this  concentration  into  the GC/MS to  determine recovery  of surrogate
standards in  all  matrix spikes.    Take  into  account  all  dilutions of  sample
extracts.

      5.9    Acetone, hexane, methylene  chloride,  isooctane,  carbon  disulfide,
toluene,  and other appropriate solvents - Pesticide quality or equivalent


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1    See  the  introductory material  to this  chapter,  Organic Analytes,
Sec. 4.1.
7.0   PROCEDURE

      7.1   Sample  preparation  -  Samples  must  be  prepared  by  one  of  the
following methods prior to GC/MS analysis.

      Matrix                              Methods
      Water                               3510, 3520
      Soil/sediment                       3540, 3541, 3550
      Waste                               3540, 3541, 3550, 3580

            7.1.1 Direct  injection  -  In   very   limited  applications  direct
      injection of the sample into  the GC/MS system with a 10 juL syringe may be
      appropriate.     The  detection   limit   is   very   high   (approximately
      10,000 /xg/L);  therefore,  it is only  permitted where  concentrations  in
      excess of  10,000 fj.g/1 are expected.   The  system must be  calibrated  by
      direct injection.
                                  8270B - 11                        Revision  2
                                                                September 1994

-------
      7.2    Extract cleanup - Extracts may be cleaned up by any of the following
methods prior to GC/MS analysis.

      Compounds                            Methods
      Phenols                              3630, 3640, 8040"
      Phthalate esters                     3610, 3620, 3640
      Nitrosamines                         3610, 3620, 3640
      Organochlorine pesticides & PCBs     3620, 3660
      Nitroaromatics and cyclic ketones    3620, 3640
      Polynuclear aromatic hydrocarbons    3611, 3630, 3640
      Haloethers                           3620, 3640
      Chlorinated hydrocarbons             3620, 3640
      Organophosphorus pesticides          3620
      Petroleum waste                      3611, 3650
      All priority pollutant base,
          neutral, and acids               3640

      8      Method 8040 includes a derivatization technique followed by GC/ECD
             analysis,   if interferences are encountered on GC/FID.

      7.3    Initial calibration - The  recommended GC/MS operating  conditions:

      Mass range:             35-500 amu
      Scan time:              1 sec/scan
      Initial temperature:    40°C,  hold for 4  minutes
      Temperature program:    40-270°C at 10°C/min
      Final  temperature:      270°C,  hold until benzo[g,h,i]perylene has eluted
      Injector temperature:   250-300°C
      Transfer line temperature: 250-300°C
      Source temperature:     According to manufacturer's specifications
      Injector:               Grob-type, splitless
      Sample volume:          1-2 juL
      Carrier gas:            Hydrogen at 50 cm/sec or helium at 30 cm/sec

      (Split injection is  allowed  if the sensitivity of the mass spectrometer
      is sufficient).

             7.3.1  Each GC/MS system must be hardware-tuned to  meet the criteria
      in Table  3  for a 50 ng  injection of DFTPP.   Analyses  should  not begin
      until  all  these  criteria are  met.    Background  subtraction  should  be
      straightforward and designed only to eliminate  column  bleed or instrument
      background ions.  The GC/MS  tuning standard  should also  be  used to assess
      GC column performance  and injection port  inertness.  Degradation of DDT
      to DDE and ODD should not exceed 20%.   (See  Sec. 8.3.1 of Method 8081 for
      the  percent  breakdown  calculation).   Benzidine and pentachlorophenol
      should be present at their  normal  responses,  and  no peak tailing should
      be visible.   If degradation is  excessive and/or  poor chromatography is
      noted, the injection port may require cleaning.   It  may  also be necessary
      to break  off  the  first 6-12 in. of the capillary column.  The  use of a
      guard  column  (Sec.  4.1.6) between the injection  port and the analytical
      column may help prolong analytical column performance.
                                  8270B - 12                        Revision 2
                                                                September 1994

-------
      7.3.2  The internal standards selected in Sec. 5.4 should permit most
of the components  of  interest  in  a chromatogram to have retention times
of 0.80-1.20 relative to one of the internal standards.  Use the base peak
ion  from  the  specific  internal  standard  as  the  primary  ion  for
quantitation (see Table 1).  If interferences are noted, use the next most
intense ion as the quantitation ion (i.e.  for l,4-dichlorobenzene-d4, use
152 m/z for quantitation).

      7.3.3  Analyze 1 /iL of each calibration standard (containing internal
standards) and tabulate the area of the primary characteristic ion against
concentration for each compound (as indicated in Table 1).   Figure 1 shows
a chromatogram of a calibration standard containing  base/neutral and acid
analytes.  Calculate response  factors  (RFs)  for  each compound  relative to
one of the internal standards as follows:

                         RF =  (AxCis)/(AisCx)

where:

      Ax     =     Area of the characteristic ion  for  the  compound being
                  measured.
      Ais     =     Area of the characteristic ion  for  the specific internal
                  standard.
      Cis     =     Concentration of the specific internal standard (ng/^L).
      Cx     =     Concentration of the compound  being  measured (ng/^L).

      7.3.4  A  system  performance  check must be  performed  to ensure that
minimum average RFs  are met  before the calibration curve  is used.   For
semivolatiles,  the  System Performance   Check  Compounds   (SPCCs)  are:
N-nitroso-di-n-propylamine;hexachlorocyclopentadiene;2,4~dinitro-phenol;
and 4-nitrophenol.  The minimum acceptable average RF for these compounds
is 0.050.  These SPCCs typically have very low RFs (0.1-0.2) and tend to
decrease in response as the chromatographic system begins  to deteriorate
or the  standard  material begins to deteriorate.   They are usually  the
first to  show  poor performance.   Therefore, they must meet the minimum
requirement when the system is calibrated.

             7.3.4.1     The percent  relative  standard deviation  (%RSD)
      should be less  than  15%  for  each compound.   However, the %RSD for
      each individual Calibration  Check Compound (CCC)  (see  Table 4) must
      be less than 30%.  The relative  retention  times of each compound in
      each calibration  run  should agree within  0.06  relative retention
      time  units.    Late-eluting  compounds  usually   have  much  better
      agreement.

               SD
      %RSD =   _   x  100
               RF
where:

      RSD    =     relative  standard deviation.
      RF     =     mean of 5 initial RFs for  a compound.
      SD     =     standard  deviation of average  RFs  for a  compound.


                           8270B  - 13                        Revision  2
                                                          September 1994

-------
      SD =
 N  (RFj - RF):
 I  	
i=l  N - 1
            where:
                  RF>    = RF for each of the 5 calibration levels
                  N      = Number of RF values (i.e.,  5)

            7.3.4.2     If the %RSD of any CCC is 30% or greater, then the
      chromatographic system is  too reactive for analysis  to begin.  Clean
      or replace the injector liner and/or capillary column,  then repeat
      the calibration procedure beginning with  section 7.3.

      7.3.5 Linearity -  If the %RSD of any compound is 15% or less,  then
the  relative   response   factor   is   assumed   to  be  constant  over  the
calibration range, and the average relative  response  factor  may  be  used
for quantitation (Sec. 7.6.2).

            7.3.5.1     If the  %RSD of any compound is greater than  15%,
      construct   calibration   curves  of   area  ratio   (A/Ajs)   versus
      concentration using first  or higher order regression fit  of the five
      calibration points.  The  analyst should select the regression order
      which introduces the least calibration error into the quantitation
      (Sec. 7.6.2.2  and  7.6.2.3).   The  use  of calibration curves  is  a
      recommended alternative to average  response factor calibration, and
      a useful diagnostic of standard preparation accuracy and absorption
      activity in the chromatographic system.

7.4   Daily GC/MS calibration

      7.4.1 Prior to analysis of samples, the GC/MS tuning standard  must
be analyzed.  A 50 ng injection of DFTPP must result  in  a mass spectrum
for DFTPP which meets the criteria given  in Table 3.  These criteria  must
be demonstrated during each 12  hour shift.

      7.4.2 A calibration standard(s)  at  mid-concentration containing all
semivolatile  analytes,   including   all   required  surrogates,  must  be
analyzed every 12  hours during analysis.  Compare the instrument response
factor from the standards  every 12 hours  with the  SPCC  (Sec.  7.4.3) and
CCC (Sec. 7.4.4) criteria.

      7.4.3 System   Performance  Check  Compounds  (SPCCs):    A  system
performance check must be made during  every 12 hour shift.  For each  SPCC
compound in the daily calibration a minimum response factor of 0.050  must
be obtained.  This is the  same  check  that  is  applied  during  the  initial
calibration.  If the  minimum response factors are not met, the system must
be evaluated,  and corrective action must be taken before  sample analysis
begins.   The minimum RF  for  semivolatile  SPCCs  is  0.050.   Some possible
problems  are   standard   mixture   degradation,  injection   port   inlet
contamination, contamination at the front end of  the  analytical  column,
                            8270B - 14                        Revision 2
                                                          September 1994

-------
and active sites  in the column or chromatographic system.  This check must
be met before analysis begins.

      7.4.4 Calibration  Check  Compounds  (CCCs):    After  the  system
performance check is  met,  CCCs  listed in Table 4 are  used  to  check the
validity of the initial  calibration.

      Calculate the percent drift using:

                              c   -  c
                                    c
                    % Drift = -  x 100
                                 C,

where:

      C,  =   Calibration Check Compound standard concentration.
      Cc =  Measured concentration using selected quantitation method.

      If the percent difference  for each CCC is less  than or equal to 20%,
the initial calibration is assumed to be valid.  If the criterion is not
met (>  20%  drift)  for  any one  CCC,  corrective  action must  be taken.
Problems similar to those  listed under SPCCs could affect this criterion.
If no  source of the problem can  be determined after corrective action has
been  taken, a  new  five-point  calibration  must  be  generated.    This
criterion must  be met before sample analysis begins.  If the CCCs are not
analytes required by the permit, then all required analytes must meet the
20% drift criterion.

      7.4.5 The  internal  standard responses  and retention times  in the
calibration check standard must  be evaluated  immediately after or during
data acquisition.  If the retention time for any internal standard changes
by more than 30  seconds from the  last  calibration  check  (12  hours), the
chromatographic system must be inspected for malfunctions and corrections
must  be  made,  as required.   If the  EICP  area for any of the  internal
standards changes by a factor of two  (-50% to +100%) from the last daily
calibration check  standard, the mass spectrometer must be  inspected for
malfunctions  and  corrections   must  be  made,  as  appropriate.    When
corrections are made, reanalysis of samples analyzed while the system was
malfunctioning is required.

7.5   GC/MS analysis

      7.5.1 It  is  highly  recommended that the extract be  screened  on  a
6C/FID or  GC/PID using the  same  type  of  capillary column.    This  will
minimize  contamination of  the  GC/MS  system  from  unexpectedly  high
concentrations  of organic  compounds.

      7.5.2 Spike  the  1 ml extract obtained from sample preparation with
10 nL of the internal  standard  solution just prior to  analysis.

      7.5.3 Analyze  the 1 mi extract by GC/MS using a 30 m x 0.25 mm (or
0.32 mm) silicone-coated fused-silica capillary column.  The volume to be
injected should ideally contain  100 ng of base/neutral  and 200 ng of acid


                            8270B  -  15                         Revision  2
                                                          September 1994

-------
surrogates  (for  a  1  p.1  injection).    The  recommended  GC/MS  operating
conditions to be used are specified in Sec.  7.3.

      7.5.4  If the response for any quantitation ion exceeds the initial
calibration curve range of the GC/MS  system,  extract  dilution  must take
place.  Additional  internal  standard must  be added to the diluted extract
to maintain  the  required  40  ng//iL   of  each  internal  standard  in  the
extracted volume.  The diluted extract must  be reanalyzed.

      7.5.5  Perform  all  qualitative  and quantitative measurements  as
described in Sec. 7.6.  Store the extracts at 4°C,  protected from  light
in screw-cap vials equipped with unpierced Teflon lined septa.

7.6   Data interpretation

      7.6.1  Qualitative analysis

             7.6.1.1     The  qualitative  identification   of   compounds
      determined  by this  method  is  based  on  retention  time,   and  on
      comparison of the sample mass spectrum, after background correction,
      with  characteristic   ions   in  a  reference mass spectrum.    The
      reference mass spectrum must be generated  by  the laboratory using
      the conditions of this  method.   The characteristic  ions  from  the
      reference mass spectrum  are defined to  be the three ions of greatest
      relative intensity,  or any  ions  over 30% relative intensity if less
      than three such  ions  occur in  the reference  spectrum.   Compounds
      should be identified as present when the criteria below are met.

                  7.6.1.1.1    The intensities  of the characteristic ions
             of a compound maximize in  the  same scan or within one scan of
             each  other.   Selection of  a  peak  by  a data  system  target
             compound  search  routine   where  the  search  is  based  on  the
             presence  of  a target  chromatographic  peak containing  ions
             specific  for  the  target  compound   at  a  compound-specific
             retention time will be accepted as meeting this criterion.

                  7.6.1.1.2    The RRT of the sample component  is  within
             ± 0.06 RRT units of the RRT of the standard component.

                  7.6.1.1.3    The   relative     intensities    of    the
             characteristic  ions   agree   within  30%   of  the   relative
             intensities  of  these  ions  in  the   reference   spectrum.
             (Example:   For  an  ion  with an  abundance  of 50%   in  the
             reference spectrum, the' corresponding abundance in  a  sample
             spectrum can range between 20% and 80%.)

                  7.6.1.1.4    Structural isomers that produce very similar
            mass spectra should  be identified as individual isomers  if
             they   have   sufficiently   different  GC   retention   times.
             Sufficient GC  resolution  is  achieved  if  the height  of  the
             valley between two isomer peaks is  less than  25% of  the  sum
             of the two peak heights.  Otherwise, structural  isomers  are
             identified as isomeric pairs.


                            8270B - 16                       Revision 2
                                                          September 1994

-------
            7.6.1.1.5    Identification  is  hampered  when  sample
      components are not  resolved  chromatographically and produce
      mass  spectra  containing ions  contributed  by more  than  one
      analyte.  When gas chromatographic peaks obviously represent
      more  than one sample  component (i.e.,  a broadened peak with
      shoulder(s)  or  a  valley  between  two  or  more  maxima),
      appropriate  selection  of  analyte  spectra  and  background
      spectra is important.   Examination  of  extracted ion current
      profiles  of  appropriate  ions  can aid  in  the  selection  of
      spectra,  and  in  qualitative  identification of  compounds.
      When analytes  coelute (i.e.,  only  one chromatographic peak is
      apparent), the identification  criteria  can  be met,  but each
      analyte spectrum will contain extraneous ions contributed by
      the coeluting compound.

      7.6.1.2     For  samples containing components not associated
with the calibration standards, a library search may be made for the
purpose of tentative identification.  The necessity to  perform this
type of  identification  will be determined by the purpose  of  the
analyses  being   conducted.    Computer  generated  library  search
routines  should  not   use  normalization  routines   that   would
misrepresent the library  or  unknown  spectra when  compared to each
other.   For  example, the RCRA permit or waste delisting  requirements
may require the  reporting  of nontarget analytes.  Only  after visual
comparison of sample spectra with the nearest library searches will
the  mass  spectral   interpretation  specialist  assign  a  tentative
identification.  Guidelines for making tentative identification are:

      (1)   Relative  intensities of  major ions  in the  reference
spectrum (ions > 10% of  the  most abundant ion) should be present in
the sample spectrum.

      (2)   The  relative intensities of the major ions  should agree
within ± 20%.  (Example:   For an ion with an abundance of 50% in the
standard spectrum,  the corresponding sample  ion  abundance must be
between 30 and 70%.)

      (3)   Molecular ions present in the reference spectrum should
be present in the sample spectrum.

      (4)   Ions  present  in  the sample  spectrum  but  not  in  the
reference  spectrum  should  be  reviewed for  possible  background
contamination or presence of coeluting compounds.

      (5)   Ions  present  in the reference spectrum but  not  in  the
sample spectrum  should be reviewed for possible subtraction from the
sample spectrum  because  of background contamination  or  coeluting
peaks.   Data system library  reduction programs can sometimes create
these discrepancies.
                      8270B -  17                         Revision  2
                                                    September  1994

-------
7.6.2 Quantitative analysis

      7.6.2.1     When  a  compound   has  been   identified,   the
quantitation  of  that  compound will  be based  on the  integrated
abundance from the EICP of the primary characteristic ion.

      7.6.2.2     If  the  %RSD of  a  compound's relative  response
factor is 15% or  less,  then the concentration in the extract may be
determined  using  the  average response  factor  (RF)  from  initial
calibration data (7.4.5.2) and the following equation:.

                      (Ax  x CJ
      C
       ex
                      (A,-,  x  RF)
where Cex  is the  concentration  of the  compound  in  the extract,  and
the other terms are as defined in Sec. 7.4.3.

      7.6.2.3     Alternatively,  the regression line fitted to  the
initial  calibration (Sec.  7.3.5.1) may be used for determination of
the extract concentration.

      7.6.2.4     Compute  the concentration of the  analyte  in  the
sample using the equations in Sees.  7.6.2.4.1 and 7.6.2.4.2.

            7.6.2.4.1    The  concentration  of  the  analyte  in  the
      liquid  phase   of   the  sample  is  calculated   using   the
      concentration of the analyte in  the extract and the volume of
      liquid extracted, as follows:

            Concentration  in  liquid  (M9/L)  =  (Cex x_V8J
                                                V0

      where:

            Vex    =     extract volume,  in  ml
            V0     =     volume  of liquid  extracted,  in L.

            7.6.2.4.2    The  concentration  of the  analyte  in  the
      solid   phase  of  the   sample   is   calculated   using   the
      concentration of the pollutant in the extract and the weight
      of the solids,  as follows:

            Concentration  in  solid (/zg/kg)  =  (C^ x _VBJ
      where:

            Vex    =     extract volume,  in ml
            Ws     =     sample weight,  in kg.

      7.6.2.5     Where applicable, an estimate of concentration for
noncalibrated components in the sample should be made.  The formulae

                      8270B  -  18                        Revision 2
                                                    September 1994

-------
            given  above  should  be  used  with the following modifications:  The
            areas Ax and A.   should be from  the  total  ion chromatograms and the
            RF  for  the  compound  should  be  assumed to be 1.  The concentration
            obtained  should  be reported indicating  (1) that the  value  is an
            estimate  and (2)  which internal  standard  was used  to determine
            concentration.     Use   the   nearest  internal   standard  free  of
            interferences.

                   7.6.2.6     Quantitation   of  multicomponent compounds  (e.g.
            Aroclors)   is  beyond   the   scope   of  Method  8270.     Normally,
            quantitation is  performed using  a GC/ECD  by Method 8081.


8.0   QUALITY CONTROL

      8.1   Each  laboratory  that uses  these methods  is required  to operate a
formal quality control program.  The minimum requirements of this program consist
of an initial  demonstration of laboratory capability and an ongoing analysis of
spiked samples  to evaluate  and  document quality  data.    The laboratory must
maintain records to document  the quality of  the data  generated.   Ongoing data
quality checks are compared with established performance criteria to determine
if the results of analyses meet the performance characteristics  of the method.
When results of sample  spikes indicate  atypical  method  performance, a quality
control reference  sample (Sec.  8.5.1)  must be  analyzed  to confirm  that  the
measurements were performed in an in-control mode of operation.

      8.2   Before  processing any  samples, the  analyst   should  demonstrate,
through the analysis of a method blank,  that interferences from the analytical
system, glassware, and reagents are under control.  Each time a  set of samples
is extracted or  there is a change  in  reagents, a method blank should be processed
as a safeguard against chronic laboratory contamination.  The blanks should be
carried through all stages of sample preparation and measurement.

      8.3   The  experience   of  the  analyst  performing  GC/MS  analyses  is
invaluable to  the success of  the  methods.  Each  day that analysis is performed,
the  daily  calibration   standard  should  be  evaluated  to determine  if  the
chromatographic system is operating properly.   Questions  that  should be asked
are:   Do  the  peaks look normal?;  Is  the response obtained comparable  to  the
response  from previous  calibrations?    Careful  examination  of the  standard
chromatogram can  indicate  whether  the column  is  still  good, the  injector is
leaking,  the injector septum needs  replacing, etc.  If any changes are made to
the system (e.g. column changed), recal ibration of the system must take place.

      8.4   Required instrument QC  is found  in  the following sections

            8.4.1  The   GC/MS  system  must  be  tuned   to  meet   the   DFTPP
      specifications in Sees. 7.3.1 and 7.4.1.

            8.4.2  There  must  be  an initial  calibration  of the  GC/MS system as
      specified in Sec. 7.3.

            8.4.3  The GC/MS system  must meet  the SPCC criteria specified in Sec.
      7.4.3 and the CCC criteria in Sec. 7.4.4, each 12 hours.


                                  8270B - 19                        Revision 2
                                                                September 1994

-------
      8.5   To  establish  the  ability  to  generate  acceptable  accuracy  and
precision, the analyst must perform the following operations.

            8.5.1 A  quality  control  (QC)   reference  sample  concentrate  is
      required containing  each  base/neutral  analyte at  a concentration of 100
      mg/L and each acid  analyte  at  a  concentration of  200 mg/L  in acetone or
      methanol.   (See Sec. 5.5.1 of Method 3500  for  minimum requirements.)  The
      QC  reference  sample concentrate  may be  prepared  from pure  standard
      materials  or  purchased  as  certified  solutions.    If  prepared  by  the
      laboratory, the QC reference sample concentrate must be made using stock
      standards prepared  independently from those used for calibration.

            8.5.2 Using a pipet, prepare QC reference samples at a concentration
      of  100 /itg/L by adding 1.00 ml of QC reference sample concentrate to each
      of  four 1-L aliquots of water.

            8.5.3 Analyze  the well-mixed QC reference samples according to the
      method beginning in  Sec. 7.1 with extraction of the samples.

            8.5.4 Calculate the average recovery (x) in  /xg/L, and the standard
      deviation of the recovery  (s) in M9/U  for each  analyte of interest using
      the four results.

            8.5.5 For  each analyte  compare s  and x with  the  corresponding
      acceptance criteria_for precision  and accuracy,  respectively,  found in
      Table 6.   If s  and  x for  all analytes  meet  the  acceptance  criteria,  the
      system performance is acceptable and analysis of actual  samples can_begin.
      If  any individual s exceeds  the precision  limit  or  any  individual x falls
      outside  the  range   for   accuracy,   then   the  system  performance  is
      unacceptable for that analyte.

            NOTE: The  large number of analytes  in Table 6 present a substantial
                  probability that one  or  more will  fail at  least  one of the
                  acceptance  criteria when  all  analytes  of a given method are
                  analyzed.

            8.5.6 When one or more of the analytes tested fail at least one of
      the  acceptance  criteria,   the  analyst must   proceed  according  to  Sec.
      8.5.6.1 or 8.5.6.2.

                  8.5.6.1     Locate  and  correct the source of the problem and
            repeat the  test for all  analytes of interest  beginning  with  Sec.
            8.5.2.

                  8.5.6.2     Beginning with Sec. 8.5.2, repeat  the  test  only
            for those analytes that failed to meet criteria.  Repeated failure,
            however, will confirm a general  problem with the measurement system.
            If this  occurs,  locate and correct the source  of  the  problem and
            repeat the test for all  compounds  of  interest  beginning with  Sec.
            8.5.2.

      8.6   The  laboratory must,  on  an ongoing  basis, analyze a  method blank,
a matrix spike, and a replicate for each analytical  batch  (up  to a maximum of 20


                                  8270B - 20                        Revision 2
                                                                September 1994

-------
samples/batch) to assess accuracy.   For soil  and waste samples where detectable
amounts of organics are present, replicate samples may be appropriate in place
of matrix  spiked  samples.   For laboratories analyzing one  to  ten samples per
month, at least one spiked sample per month is required.

            8.6.1  The  concentration  of the  spike  in  the  sample  should  be
      determined as follows:

                   8.6.1.1     If, as in compliance monitoring, the concentration
            of  a  specific analyte  in the sample  is being checked  against  a
            regulatory  concentration  limit,  the  spike  should  be at that limit
            or  1 to 5 times higher  than  the background concentration determined
            in  Sec. 8.6.2, whichever  concentration would be larger.

                   8.6.1.2     If the  concentration of a  specific  analyte in  a
            water  sample is not being checked against a limit specific to that
            analyte, the spike should be at 100 ^g/L or 1  to 5 times higher than
            the  background concentration determined in  Step  8.6.2,  whichever
            concentration  would be larger.   For other  matrices,  recommended
            spiking concentration  is  20  times the EQL.

                   8.6.1.3     If  it  is  impractical  to  determine  background
            levels before spiking (e.g. maximum holding  times will be exceeded),
            the   spike   concentration   should   be   at  (1)   the   regulatory
            concentration  limit, if any; or,  if  none (2)  the  larger  of either
            5 times  higher than the  expected background concentration  or 100
            M9/L.  For  other matrices, recommended spiking concentration is 20
            times  the EQL.

            8.6.2  Analyze  one  sample  aliquot   to   determine  the  background
      concentration  (B)  of  each  analyte.   If  necessary,  prepare  a new  QC
      reference sample concentrate  (Sec. 8.5.1)  appropriate for the background
      concentration in the sample.   Spike a second sample aliquot  with 1.00 ml
      of the  QC  reference  sample concentrate and analyze  it  to determine the
      concentration after  spiking  (A) of each analyte.   Calculate  each percent
      recovery (p) as 100(A-B)%/T,  where  T is the  known true  value of the spike.

            8.6.3  Compare  the percent recovery (p) for each analyte in a water
      sample with  the  corresponding QC acceptance criteria found  in  Table 6.
      These acceptance  criteria were calculated  to  include an allowance for
      error in  measurement of  both  the background  and  spike  concentrations,
      assuming a spike  to background ratio of 5:1.  This error will  be  accounted
      for to the extent  that  the analyst's spike to background ratio approaches
      5:1.   If spiking was  performed at a concentration lower than 100  /zg/L, the
      analyst must use either the QC acceptance  criteria  presented in Table 6,
      or optional  QC acceptance  criteria  calculated  for  the  specific  spike
      concentration.  To calculate optional  acceptance criteria for the recovery
      of an analyte:  (1)  Calculate accuracy (x') using the equation  found in
      Table 7,  substituting  the spike concentration  (T)  for C; (2)  calculate
      overall  precision (S')  using  the equation in Table 7,  substituting x' for
      x; (3)  calculate  the range  for  recovery  at the spike  concentration  as
      (100x'/T)  ± 2.44(100S'/T)%.
                                  8270B - 21                        Revision 2
                                                                September 1994

-------
            8.6.4  If any individual p falls  outside  the designated range for
      recovery,  that analyte  has  failed  the acceptance  criteria.   A check
      standard containing each analyte  that failed the criteria must be analyzed
      as described  in Sec. 8.7.

      8.7   If  any  analyte  in  a  sample  fails  the  acceptance  criteria  for
recovery in Sec.  8.6, a  QC reference sample containing each analyte that failed
must be prepared and analyzed.

      NOTE: The  frequency for the  required analysis of  a  QC reference sample
            will depend upon the number of analytes  being  simultaneously tested,
            the  complexity of the  sample  matrix,  and the performance  of the
            laboratory.   If the  entire  list of analytes  in Table  6  must be
            measured in  the sample  in  Sec.  8.6,  the  probability that  the
            analysis of a QC reference  sample  will be required is high.   In this
            case, the QC reference sample should  be  routinely analyzed with the
            spiked  sample.

            8.7.1  Prepare the QC reference sample  by adding 1.0  mi of the QC
      reference sample concentrate  (Sec. 8.5.1 or 8.6.2) to  1 L of water.  The
      QC  reference  sample  needs  only to  contain  the  analytes  that  failed
      criteria in the test in Sec.  8.6.

            8.7.2  Analyze the QC reference sample to determine the concentration
      measured  (A)  of each analyte.   Calculate  each  percent recovery  (pj as
      100(A/T)%, where T  is  the true value of the standard concentration.

            8.7.3  Compare the percent  recovery (ps)  for  each analyte with the
      corresponding  QC acceptance criteria found  in Table 6.   Only analytes that
      failed the test in  Sec. 8.6 need to be compared with these criteria.  If
      the recovery of any such analyte falls outside the designated range, the
      laboratory performance  for  that  analyte is judged  to  be out of control,
      and  the problem  must  be   immediately  identified  and  corrected.   The
      analytical result for that  analyte in the unspiked sample is suspect and
      may not be reported for regulatory compliance purposes.

      8.8   As part  of  the  QC program  for  the laboratory,  method accuracy for
each matrix studied must be assessed and  records  must be maintained.  After the
analysis of five spiked samplesjof the same matrix) as in Sec. 8.6, calculate
the  average  percent recovery  (p)  and  the standard deviation of  the  percent
recovery (sp).   Express  the accuracy assessment as  a percent  recovery interval
from p  - 2sp to p  +  2sp.   If p = 90% and sp  = 10%,  for  example,  the accuracy
interval is  expressed as  70-110%.   Update  the  accuracy assessment for  each
analyte  on  a  regular  basis  (e.g.  after  each  five   to   ten  new  accuracy
measurements).

      8.9   The following procedure should be  performed to determine acceptable
accuracy and precision limits for surrogate standards.

            8.9.1  For each  sample analyzed,  calculate  the  percent recovery of
      each surrogate in the sample.
                                  8270B - 22                        Revision 2
                                                                September 1994

-------
             8.9.2  Once a minimum of thirty samples of the same matrix have been
      analyzed,  calculate  the  average  percent  recovery  (P)  and  standard
      deviation of the  percent recovery  (s) for  each of  the  surrogates.

             8.9.3  For a given matrix,  calculate the upper  and  lower control
      limit  for method performance for each surrogate standard.  This should be
      done as follows:

             Upper  Control  Limit  (UCL)  =  P  + 3s
             Lower  Control  Limit  (LCL)  =  P  - 3s

             8.9.4  For aqueous and soil matrices, these  laboratory-established
      surrogate  control limits  should,  if applicable,  be  compared  with  the
      control limits listed in Table 8.  The limits  given in  Table 8 are multi-
      laboratory  performance-based  limits for  soil  and  aqueous  samples,  and
      therefore,  the  single-laboratory limits established in Sec.  8.9.3 must
      fall within those given in Table 8 for these matrices.

             8.9.5  If  recovery is not within limits, the following procedures are
      required.

             •      Check to  be  sure   there  are  no  errors  in  calculations,
                   surrogate  solutions  and  internal  standards.   Also,  check
                   instrument  performance.

             •      Recalculate  the data and/or reanalyze the  extract if any of
                   the above checks  reveal  a problem.

             •      Reextract and  reanalyze  the  sample if none of  the above  are
                   a problem or flag the  data  as  "estimated concentration".

             8.9.6  At  a minimum, each laboratory should update  surrogate recovery
      limits on a matrix-by-matrix basis, annually.

      8.10   It  is recommended  that the  laboratory adopt additional  quality
assurance practices for use with this method.   The specific practices that  are
most productive depend  upon the needs  of the  laboratory  and  the  nature of  the
samples.   Field duplicates may  be analyzed  to  assess  the  precision  of  the
environmental measurements. When doubt exists  over the  identification of a peak
on the chromatogram,  confirmatory techniques such as gas chromatography with a
dissimilar column, specific element  detector,  or a  mass  spectrometer  must  be
used.   Whenever possible,  the  laboratory  should analyze standard  reference
materials and participate in relevant performance evaluation  studies.


9.0   METHOD PERFORMANCE

      9.1    Method 8250  (the  packed column  version  of  Method 8270)  was tested
by 15 laboratories using  organic-free reagent water, drinking water,  surface
water, and industrial  wastewaters spiked at six concentrations over the range 5-
1,300 /Ltg/L.  Single operator  accuracy  and  precision,  and method  accuracy were
found to be directly related to the concentration of  the analyte and essentially
                                  8270B - 23                        Revision 2
                                                                September 1994

-------
independent  of  the   sample   matrix.   Linear  equations  to  describe  these
relationships are presented in Table 7.

      9.2    Chromatograms from calibration standards analyzed with Day 0 and Day
7  samples  were  compared to detect  possible  deterioration  of  GC performance.
These recoveries (using Method 3510 extraction) are presented in Table  9.

      9.3    Method  performance  data  (using  Method  3541  Automated   Soxhlet
extraction) are  presented in Table 10.  Single laboratory accuracy and precision
data were  obtained  for semivolatile organics  in  a clay soil by  spiking at a
concentration of 6 mg/kg for each compound.  The spiking solution was mixed into
the soil during  addition  and then  allowed  to  equilibrate for approximately 1 hr
prior to extraction.   The spiked samples were then extracted  by  Method 3541
(Automated Soxhlet).  Three determinations were performed and each extract was
analyzed by gas chromatography/ mass  spectrometry following Method 8270.   The low
recovery of the more volatile compounds is probably due  to volatilization losses
during equilibration.  These  data are listed  in Table  11  and were taken from
Reference 9.
10.0  REFERENCES

1.    U.S. EPA 40 CFR  Part 136, "Guidelines Establishing Test  Procedures for the
      Analysis of Pollutants Under the Clean Water Act,  Method 625," October 26,
      1984.

2.    U.S.  EPA  Contract  Laboratory Program,  Statement of  Work  for  Organic
      Analysis, July  1985, Revision.

3.    Eichelberger, J.W., L.E. Harris,  and  W.L.  Budde, "Reference Compound to
      Calibrate   Ion   Abundance   Measurement   in   Gas   Chromatography-Mass
      Spectrometry Systems," Analytical Chemistry, 47, 995-1000,  1975.

4.    "Method Detection Limit for Methods  624  and  625," Olynyk, P., W.L. Budde,
      and J.W. Eichelberger, Unpublished report, October 1980.

5.    "Inter!aboratory Method Study  for EPA Method 625-Base/Neutrals, Acids, and
      Pesticides," Final Report for EPA Contract 68-03-3102  (in preparation).

6.    Burke,  J.A.  "Gas  Chromatography  for Pesticide  Residue Analysis;  Some
      Practical  Aspects,"  Journal  of  the  Association of  Official  Analytical
      Chemists, 48, 1037, 1965.

7.    Lucas, S.V.; Kornfeld,  R.A. "GC-MS  Suitability  Testing of RCRA Appendix
      VIII and Michigan List Analytes "; U.S.  Environmental Protection Agency,
      Environmental Monitoring and  Support Laboratory, Cincinnati,  OH 45268,
      February 20, 1987, Contract No. 68-03-3224.

8.    Engel, T.M.; Kornfeld,  R.A.;  Warner,  J.S.;  Andrews,  K.D.   "Screening of
      Semivolatile Organic Compounds for  Extractability  and  Aqueous  Stability
      by  SW-846,     Method   3510";  U.S.   Environmental   Protection  Agency,
      Environmental Monitoring and  Support Laboratory, Cincinnati,  OH 45268,
      June 5, 1987, Contract 68-03-3224.


                                  8270B -  24                        Revision 2
                                                                September 1994

-------
9.     Lopez-Avila, V.  (W. Beckert,  Project  Officer);  "Development of a Soxtec
      Extraction Procedure for Extraction of Organic  Compounds from Soils and
      Sediments";    U.S.  Environmental  Protection  Agency.    Environmental
      Monitoring  and  Support Laboratory.    Las  Vegas,  NV, October  1991;  EPA
      600/X-91/140.
                                  8270B - 25                        Revision 2
                                                                September 1994

-------
                                   TABLE 1.
                CHARACTERISTIC IONS FOR SEMIVOLATILE COMPOUNDS
Compound
Retention    Primary   Secondary
Time (min.)  Ion       Ion(s)
2-Picoline
Aniline
Phenol
Bis(2-chloroethy1) ether
2-Chlorophenol
1,3-Dichlorobenzene
l,4-Dichlorobenzene-d4 (I.S.)
1,4-Dichlorobenzene
Benzyl alcohol
1,2-Dichlorobenzene
N-Nitrosomethylethyl amine
Bis(2-chloroisopropyl) ether
Ethyl carbamate
Thiophenol (Benzenethiol)
Methyl methanesulfonate
N-Nitrosodi-n-propylamine
Hexachloroethane
Maleic anhydride
Nitrobenzene
Isophorone
N-Nitrosodiethyl amine
2-Nitrophenol
2,4-Dimethylphenol
p-Benzoquinone
Bis(2-chloroethoxy)methane
Benzoic acid
2,4-Dichlorophenol
Trimethyl phosphate
Ethyl methanesulfonate
1,2,4-Trichlorobenzene
Naphthalene-d8 (I.S.)
Naphthalene
Hexachlorobutadiene
Tetraethyl pyrophosphate
Diethyl sulfate
4-Chloro-3-methylphenol
2-Methylnaphthalene
2-Methylphenol
Hexachloropropene
Hexachlorocyclopentadiene
N-Nitrosopyrrolidine
Acetophenone
4-Methylphenol
2,4,6-Trichlorophenol
o-Toluidine
3-Methylphenol
     3.75"      93      66,92
     5.68      93      66,65
     5.77      94      65,66
     5.82      93      63,95
     5.97     128      64,130
     6.27     146      148,111
     6.35     152      150,115
     6.40     146      148,111
     6.78     108      79,77
     6.85     146      148,111
     6.97      88      42,88,43,56
     7.22      45      77,121
     7.27      62      62,44,45,74
     7.42     110      110,66,109,84
     7.48      80      80,79,65,95
     7.55      70      42,101,130
     7.65     117      201,199
     7.65      54      54,98,53,44
     7.87      77      123,65
     8.53      82      95,138
     8.70     102      102,42,57,44,56
     8.75     139      109,65
     9.03     122      107,121
     9.13     108      54,108,82,80
     9.23      93      95,123
     9.38     122      105,77
     9.48     162      164,98
     9.53     110      110,79,95,109,140
     9.62      79      79,109,97,45,65
     9.67     180      182,145
     9.75     136      68
     9.82     128      129,127
    10.43     225      223,227
    11.07      99      99,155,127,81,109
    11.37     139      139,45,59,99,111,125
    11.68     107      144,142
    11.87     142      141
    12.40     107      107,108,77,79,90
    12.45     213      213,211,215,117,106,141
    12.60     237      235,272
    12.65     100      100,41,42,68,69
    12.67     105      71,105,51,120
    12.82     107      107,108,77,79,90
    12.85     196      198,200
    12.87     106      106,107,77,51,79
    12.93     107      107,108,77,79,90
                                  8270B - 26
                                  Revision  2
                             September  1994

-------
                                   TABLE 1.
                                  (Continued)
Compound
 Retention
 Time (min.)
Primary
Ion
Secondary
Ion(s)
2-Chloronaphthalene
N-Nitrosopiperi dine
1,4-Phenylenediamine
1-Chioronaphthalene
2-Nitroaniline
5-Chloro-2-methyl aniline
Dimethyl phthalate
Acenaphthylene
2,6-Dinitrotoluene
Phthalic anhydride
o-Anisidine
3-Nitroaniline
Acenaphthene-d10 (I.S.)
Acenaphthene
2,4-Dinitrophenol
2,6-Dinitrophenol
4-Chloroaniline
Isosafrole
Dibenzofuran
2,4-Diaminotoluene
2,4-Dinitrotoluene
4-Nitrophenol
2-Naphthylamine
1,4-Naphthoquinone
p-Cresidine
Dichlorovos
Diethyl phthalate
Fluorene
2,4,5-Trimethyl aniline
N-Nitrosodibutylamine
4-Chlorophenyl phenyl ether
Hydroquinone
4,6-Dinitro-2-methylphenol
Resorcinol
N-Nitrosodiphenylamine
Safrole
Hexamethyl phosphoramide
3-(Chloromethyl)pyridine hydrochl
Diphenylamine
1,2,4,5-Tetrachlorobenzene
1-Naphthylamine
1-Acetyl-2-thiourea
4-Bromophenyl phenyl ether
Toluene diisocyanate
2,4,5-Trichlorophenol
Hexachlorobenzene
     13.30      162      127,164
     13.55      114      42,114,55,56,41
     13.62      108      108,80,53,54,52
     13.65a     162      127,164
     13.75      65      92,138
     14.28      106      106,141,140,77,89
     14.48      163      194,164
     14.57      152      151,153
     14.62      165      63,89
     14.62      104      104,76,50,148
     15.00      108      80,108,123,52
     15.02      138      108,92
     15.05      164      162,160
     15.13      154      153,152
     15.35      184      63,154
     15.47      162      162,164,126,98,63
     15.50      127      127,129,65,92
     15.60      162      162,131,104,77,51
     15.63      168      139
     15.78      121      121,122,94,77,104
     15.80      165      63,89
     15.80      139      109,65
     16.00a     143      115,116
     16.23      158      158,104,102,76,50,130
     16.45      122      122,94,137,77,93
     16.48      109      109,185,79,145
     16.70      149      177,150
     16.70      166      165,167
     16.70      120      120,135,134,91,77
     16.73      84      84,57,41,116,158
     16.78      204      206,141
     16.93      110      110,81,53,55
     17.05      198      51,105
     17.13      110      110,81,82,53,69
     17.17      169      168,167
     17.23      162      162,162,104,77,103,135
     17.33      135      135,44,179,92,42
oride!7.50      92      92,127,129,65,39
     17.54a     169      168,167
     17.97      216      216,214,179,108,143,218
     18.20      143      143,115,89,63
     18.22      118      43,118,42,76
     18.27      248      250,141
     18.42      174      174,145,173,146,132,91
     18.47      196      196,198,97,132,99
     18.65      284      142,249
                                  8270B - 27
                                   Revision 2
                               September 1994

-------
                                   TABLE 1.
                                  (Continued)
Compound
Retention
Time (min.)
Primary
Ion
Secondary
Ion(s)
Nicotine
Pentachlorophenol
5-Nitro-o-toluidine
Thionazine
4-Nitroaniline
Phenanthrene-d10(i .s.)
Phenanthrene
Anthracene
1,4-Dinitrobenzene
Mevinphos
Naled
1,3-Dinitrobenzene
Diallate (cis or trans)
1,2-Dinitrobenzene
Diallate (trans or cis)
Pentachlorobenzene
5-Nitro-o-anisidine
Pentachloronitrobenzene
4-Nitroquinoline-l-oxide
Di-n-butyl phthalate
2,3,4,6-Tetrachlorophenol
Dihydrosaffrole
Demeton-0
Fluoranthene
1,3,5-Trinitrobenzene
Dicrotophos
Benzidine
Trifluralin
Bromoxynil
Pyrene
Monocrotophos
Phorate
Sulfall ate
Demeton-S
Phenacetin
Dimethoate
Phenobarbital
Carbofuran
Octamethyl pyrophosphoramide
4-Aminobiphenyl
Dioxathion
Terbufos
a,a-Dimethylphenylamine
Pronamide
Aminoazobenzene
Dichlone
    18.70      84      84,133,161,162
    19.25     266      264,268
    19.27     152     77,152,79,106,94
    19.35     107     96,107,97,143,79,68
    19.37     138     138,65,108,92,80,39
    19.55     188     94,80
    19.62     178     179,176
    19.77     178     176,179
    19.83     168     168,75,50,76,92,122
    19.90     127     127,192,109,67,164
    20.03     109     109,145,147,301,79,189
    20.18     168     168,76,50,75,92,122
    20.57      86     86,234,43,70
    20.58     168     168,50,63,74
    20.78      86     86,234,43,70
    21.35     250     250,252,108,248,215,254
    21.50     168     168,79,52,138,153,77
    21.72     237     237,142,214,249,295,265
    21.73     174     174,101,128,75,116
    21.78     149     150,104
    21.88     232     232,131,230,166,234,168
    22.42     135     135,64,77
    22.72      88     88,89,60,61,115,171
    23.33     202     101,203
    23.68      75     75,74,213,120,91,63
    23.82     127     127,67,72,109,193,237
    23.87     184     92,185
    23.88     306     306,43,264,41,290
    23.90     277     277,279,88,275,168
    24.02     202     200,203
    24.08     127     127,192,67,97,109
    24.10      75     75,121,97,93,260
    24.23     188     188,88,72,60,44
    24.30      88     88,60,81,89,114,115
    24.33     108     180,179,109,137,80
    24.70      87     87,93,125,143,229
    24.70     204     204,117,232,146,161
    24.90     164     164,149,131,122
    24.95     135     135,44,199,286,153,243
    25.08     169     169,168,170,115
    25.25      97     97,125,270,153
    25.35     231     231,57,97,153,103
    25.43      58     58,91,65,134,42
    25.48     173     173,175,145,109,147
    25.72     197     92,197,120,65,77
    25.77     191     191,163,226,228,135,193
                                  8270B - 28
                                  Revision  2
                              September 1994

-------
                                   TABLE 1.
                                  (Continued)
Compound
Retention
Time (min.)
Primary
Ion
Secondary
Ion(s)
Dinoseb
Disulfoton
Fluchloralin
Mexacarbate
4,4'-Oxydianiline
Butyl benzyl phthalate
4-Nitrobiphenyl
Phosphamidon
2-Cyclohexyl-4,6-Dinitrophenol
Methyl parathion
Carbaryl
Dimethyl aminoazobenzene
Propylthiouracil
Benz(a)anthracene
Chrysene-d12 (I.S.)
3,3'-Dichlorobenzidine
Chrysene
Malathion
Kepone
Fenthion
Parathion
Anilazine
Bis(2-ethylhexyl) phthalate
3,3'-Dimethylbenzidine
Carbophenothion
5-Nitroacenaphthene
Methapyrilene
Isodrin
Captan
Chlorfenvinphos
Crotoxyphos
Phosmet
EPN
Tetrachlorvinphos
Di-n-octyl phthalate
2-Aminoanthraquinone
Barban
Aramite
Benzo(b)fluoranthene
Nitrofen
Benzo(k)fluoranthene
Chiorobenzilate
Fensulfothion
Ethion
Diethylstilbestrol
Famphur
    25.83     211     211,163,147,117,240
    25.83      88     88,97,89,142,186
    25.88     306     306,63,326,328,264,65
    26.02     165     165,150,134,164,222
    26.08     200     200,108,171,80,65
    26.43     149     91,206
    26.55     199     199,152,141,169,151
    26.85     127     127,264,72,109,138
    26.87     231     231,185,41,193,266
    27.03     109     109,125,263,79,93
    27.17     144     144,115,116,201
    27.50     225     225,120,77,105,148,42
    27.68     170     170,142,114,83
    27.83     228     229,226
    27.88     240     120,236
    27.88     252     254,126
    27.97     228     226,229
    28.08     173     173,125,127,93,158
    28.18     272     272,274,237,178,143,270
    28.37     278     278,125,109,169,153
    28.40     109     109,97,291,139,155
    28.47     239     239,241,143,178,89
    28.47     149     167,279
    28.55     212     212,106,196,180
    28.58     157     157,97,121,342,159,199
    28.73     199     199,152,169,141,115
    28.77      97     97,50,191,71
    28.95     193     193,66,195,263,265,147
    29.47      79     79,149,77,119,117
    29.53     267     267,269,323,325,295
    29.73     127     127,105,193,166
    30.03     160     160,77,93,317,76
    30.11     157     157,169,185,141,323
    30.27     329     109,329,331,79,333
    30.48     149     167,43
    30.63     223     223,167,195
    30.83     222     222,51,87,224,257,153
    30.92     185     185,191,319,334,197,321
    31.45     252     253,125
    31.48     283     283,285,202,139,253
    31.55     252     253,125
    31.77     251     251,139,253,111,141
    31.87     293     293,97,308,125,292
    32.08     231     231,97,153,125,121
    32.15     268     268,145,107,239,121,159
    32.67     218     218,125,93,109,217
                                  8270B - 29
                                  Revision 2
                              September 1994

-------
                                   TABLE 1.
                                  (Continued)
Compound
Retention
Time (min.)
Primary   Secondary
Ion       Ion(s)
Tri-p-tolyl phosphate6                32.75      368
Benzo(a)pyrene                        32.80      252
Perylene-d12  (I.S.)                    33.05      264
7,12-Dimethylbenz(a)anthracene        33.25      256
5,5-Diphenylhydantoin                 33.40      180
Captafol                              33.47       79
Dinocap                               33.47       69
Methoxychlor                          33.55      227
2-Acetylaminofluorene                 33.58      181
4,4'-Methylenebis(2-chloroani1ine)    34.38      231
3,3'-Dimethoxybenzidine               34.47      244
3-Methylcholanthrene                  35.07      268
Phosalone                             35.23      182
Azinphos-methyl                       35.25      160
Leptophos                             35.28      171
Mirex                                 35.43      272
Tris(2,3-dibromopropy1) phosphate     35.68      201
Dibenz(a,j)acridine                   36.40      279
Mestranol                             36.48      277
Coumaphos                             37.08      362
Indeno(l,2,3-cd)pyrene                39.52      276
Dibenz(a,hjanthracene                 39.82      278
Benzo(g,h,i)perylene                  41.43      276
l,2:4,5-Dibenzopyrene                 41.60      302
Strychnine                            45.15      334
Piperonyl sulfoxide                   46.43      162
Hexachlorophene                       47.98      196
Aldrin                                 --         66
Aroclor-1016                           --        222
Aroclor-1221                           --        190
Aroclor-1232                           --        190
Aroclor-1242                           --        222
Aroclor-1248                           --        292
Aroclor-1254                           --        292
Aroclor-1260                           --        360
a-BHC                                  --        183
/3-BHC                                  --        181

-------
                                   TABLE 1.
                                  (Continued)
                                  Retention     Primary    Secondary
Compound                          Time  (min.)   Ion        Ion(s)
Endosulfan sulfate                      --        272     387,422
Endrin                                  --        263     82,81
Endrin aldehyde                         --         67     345,250
Endrin ketone                           --        317     67,319
2-Fluorobiphenyl (surr.)                --        172     171
2-Fluorophenol  (surr.)                  --        112     64
Heptachlor                              --        100     272,274
Heptachlor epoxide                      --        353     355,351
Nitrobenzene-d5 (surr.)                  --         82     128,54
N-Nitrosodimethylamine                  --         42     74,44
Phenol-d6 (surr.)                        --         99     42,71
Terphenyl-d14 (surr.)                    --        244     122,212
2,4,6-Tribromophenol (surr.)            --        330     332,141
Toxaphene                               --        159     231,233
I.S.  = internal standard.
surr. = surrogate.
8Estimated retention times.
bSubstitute for the non-specific mixture,  tricresyl  phosphate.
                                  82708 - 31                        Revision 2
                                                                September 1994

-------
                           TABLE 2.
ESTIMATED QUANTITATION LIMITS (EQLs) FOR SEMIVOLATILE ORGANICS

                                       Estimated
                                      Quantitation
                                          Limits"
Ground water
Semivolatiles M9/L
Acenaphthene
Acenaphthylene
Acetophenone
2-Acetyl ami nof 1 uorene
l-Acetyl-2-thiourea
2-Aminoanthraquinone
Aminoazobenzene
4-Aminobiphenyl
Anilazine
o-Anisidine
Anthracene
Aramite
Azinphos-methyl
Barban
Benz(a)anthracene
Benzo(b)fluoranthene
Benzo(k)fluoranthene
Benzoic acid
Benzo (g , h , i ) peryl ene
Benzo(a)pyrene
p-Benzoquinone
Benzyl alcohol
Bis(2-chloroethoxy)methane
Bis(2-chloroethyl) ether
Bis(2-chloroisopropyl) ether
4-bromophenyl phenyl ether
Bromoxynil
Butyl benzyl phthalate
Captafol
Captan
Carbaryl
Carbofuran
Carbophenothion
Chlorfenvinphos
4-Chloroaniline
Chlorobenzilate
5- Chi oro- 2 -methyl aniline
4- Chi oro-3-methyl phenol
3-(Chloromethyl)pyridine hydrochloride
2-Chloronaphthalene
2-Chlorophenol
4-Chlorophenyl phenyl ether
Chrysene
Coumaphos
10
10
10
20
1000
20
10
20
100
10
10
20
100
200
10
10
10
50
10
10
10
20
10
10
10
10
10
10
20
50
10
10
10
20
20
10
10
20
100
10
10
10
10
40
Low Soil/Sediment"
M9/kg
660
660
ND
ND
ND
ND
ND
ND
ND
ND
660
ND
ND
ND
660
660
660
3300
660
660
ND
1300
660
660
660
660
NO
660
ND
ND
ND
ND
ND
ND
1300
ND
ND
1300
ND
660
660
660
660
ND
                          8270B - 32                        Revision 2
                                                        September 1994

-------






Semivolatiles
p-Cresidine
Crotoxyphos
2-Cyclohexyl-4,6-dinitrophenol
Demeton-0
Demeton-S
Diallate (cis or trans)
Diallate (trans or cis)
2,4-Diaminotoluene
Dibenz(a, j)acridine
Dibenz( a, h) anthracene
Dibenzofuran
Dibenzo(a,e)pyrene
Di-n-butyl phthalate
Dichlone
1 , 2-Di chl orobenzene
1 ,3-Di chl orobenzene
1 , 4-Di chl orobenzene
3,3'-Dichlorobenzidine
2,4-Dichlorophenol
2,6-Dichlorophenol
Dichlorovos
Dicrotophos
Diethyl phthalate
Diethylstilbestrol
Diethyl sulfate
Dimethoate
3,3'-Dimethoxybenzidine
Dimethyl aminoazobenzene
7, 12-Dimethylbenz(a)anthracene
3,3'-Dimethylbenzidine
a, a-Di methyl phenethyl amine
2,4-Dimethylphenol
Dimethyl phthalate
1,2-Dinitrobenzene
1,3-Dinitrobenzene
1,4-Dinitrobenzene
4, 6-Dinitro-2-methyl phenol
2,4-Dinitrophenol
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Dinocap
Dinoseb
5,5-Diphenylhydantoin
Di-n-octyl phthalate
TABLE 2.
(Continued)



Ground
M9/1
10
20
100
10
10
10
10
20
10
10
10
10
10
NA
10
10
10
20
10
10
10
10
10
20
100
20
100
10
10
10
ND
10
10
40
20
40
50
50
10
10
100
20
20
10


Estimated
Quantitation
Limits"
water Low Soil/Sediment"
Mg/kg
ND
ND
ND
ND
ND
ND
ND
ND
ND
660
660
ND
ND
ND
660
660
660
1300
660
ND
ND
ND
660
ND
ND
ND
ND
ND
ND
ND
ND
660
660
ND
ND
ND
3300
3300
660
660
ND
ND
ND
660
8270B - 33
    Revision 2
September 1994

-------






Semivolatiles
Disulfoton
EPN
Ethion
Ethyl carbamate
Bis(Z-ethylhexyl) phthalate
Ethyl methanesulfonate
Famphur
Fensulfothion
Fenthion
Fluchloralin
Fluoranthene
Fluorene
Hexachlorobenzene
Hexachlorobutadiene
Hexachlorocyclopentadiene
Hexachloroethane
Hexachlorophene
Hexachl oropropene
Hexamethyl phosphorami de
Hydroquinone
Indeno(l,2,3-cd)pyrene
Isodrin
Isophorone
Isosafrole
Kepone
Leptophos
Malathion
Maleic anhydride
Mestranol
Methapyrilene
Methoxychlor
3-Methyl chol anthrene
4,4'-Methylenebis(2-chloroanil
Methyl methanesulfonate
2 -Methyl naphthalene
Methyl parathion
2-Methyl phenol
3-Methylphenol
4-Methylphenol
Mevinphos
Mexacarbate
Mi rex
Monocrotophos
Naled
TABLE 2.
(Continued)
Estimated
Quantitation
Limits"
Ground water Low Soi
M9/L
10
10
10
50
10
20
20
40
10
20
10
10
10
10
10
10
50
10
20
ND
10
20
10
10
20
10
50
NA
20
100
10
10
ine) NA
10
10
10
10
10
10
10
20
10
40
20





l/Sedimentb
M9/kg
ND
ND
ND
ND
660
ND
ND
ND
ND
ND
660
660
660
660
660
660
ND
ND
ND
ND
660
ND
660
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
660
ND
660
ND
660
ND
ND
ND
ND
ND
8270B - 34                        Revision 2
                              September 1994

-------






Semivolatiles
Naphthalene
1,4-Naphthoquinone
1-Naphthylamine
2-Naphthylamine
Nicotine
5-Nitroacenaphthene
2-Nitroaniline
3-Nitroaniline
4-Nitroaniline
5-Nitro-o-anisidine
Nitrobenzene
4-Nitrobiphenyl
Nitrofen
2-Nitrophenol
4-Nitrophenol
5-Nitro-o-toluidine
4-Nitroquinoline-l -oxide
N-Nitrosodibutylamine
N-Nitrosodiethyl amine
N-Nitrosodiphenylamine
N-Nitroso-di -n-propylamine
N-Nitrosopiperidine
N-Nitrosopyrrol idine
Octamethyl pyrophosphoramide
4,4' -Oxydianil ine
Parathion
Pentachl orobenzene
Pentachl oronitrobenzene
Pentachl orophenol
Phenacetin
Phenanthrene
Phenobarbital
Phenol
1 , 4- Phenyl enedi ami ne
Phorate
Phosalone
Phosmet
Phosphamidon
Phthalic anhydride
2-Picoline
Piperonyl sulfoxide
Pronamide
Propylthiouracil
Pyrene
TABLE 2.
(Continued)
Estimated
Quantitation
Limits8
Ground water Low Soi
M9A
10
10
10
10
20
10
50
50
20
10
10
10
20
10
50
10
40
10
20
10
10
20
40
200
20
10
10
20
50
20
10
10
10
10
10
100
40
100
100
NO
100
10
100
10





l/Sedimentb
/xg/kg
660
ND
ND
ND
ND
ND
3300
3300
ND
ND
660
ND
ND
660
3300
ND
ND
ND
ND
660
660
ND
ND
ND
ND
ND
ND
ND
3300
ND
660
ND
660
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
660
8270B - 35
    Revision 2
September 1994

-------
                                   TABLE 2.
                                  (Continued)
                                               Estimated
                                              Quantitation
                                                 Limits'
                                      Ground water   Low Soil/Sedimervr
Semivolatiles                            /xg/L
Pyridine
Resorcinol
Safrole
Strychnine
Sul fall ate
Terbufos
1 , 2, 4, 5-Tetrachl orobenzene
2,3,4,6-Tetrachlorophenol
Tetrachlorvinphos
Tetraethyl pyrophosphate
Thionazine
Thiophenol (Benzenethiol)
Toluene diisocyanate
o-Toluidine
1, 2, 4-Trichl orobenzene
2,4,5-Trichlorophenol
2 , 4 , 6-Tr i chl orophenol
Trifluralin
2, 4, 5 -Tri methyl anil ine
Trimethyl phosphate
1,3,5-Trinitrobenzene
Tris(2,3-dibromopropyl ) phosphate
Tri-p-tolyl phosphate(h)
0,0,0-Triethyl phosphorothioate
ND
100
10
40
10
20
10
10
20
40
20
20
100
10
10
10
10
10
10
10
10
200
10
NT
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
660
660
660
ND
ND
ND
ND
ND
ND
ND
a Sample EQLs are highly matrix-dependent.  The EQLs  listed herein are provided
  for guidance and may not always be achievable.
b EQLs  listed  for soil/sediment are based  on  wet weight.   Normally data are
  reported on a  dry  weight basis, therefore,  EQLs will  be higher based on the
  % dry weight of each  sample.   These  EQLs  are based on a 30 g sample and gel
  permeation chromatography cleanup.
ND = Not determined.
NA = Not applicable.
NT = Not tested.
Other Matrices                                       Factor0

High-concentration soil and sludges by sonicator775
Non-water miscible waste                             75

°EQL  =  (EQL  for Low Soil/Sediment  given  above in Table 2) X (Factor).


                                  8270B - 36                        Revision 2
                                                                September 1994

-------
                                   TABLE 3.
                  DFTPP KEY  IONS  AND  ION ABUNDANCE CRITERIA"'6
       Mass
Ion Abundance Criteria
       51

       68
       70

      127

      197
      198
      199

      275

      365

      441
      442
      443
30-60% of mass 198

< 2% of mass 69
< 2% of mass 69

40-60% of mass 198

< 1% of mass 198
Base peak, 100% relative abundance
5-9% of mass 198

10-30% of mass 198

> 1% of mass 198

Present but less than mass 443
> 40% of mass 198
17-23% of mass 442
   a  See Reference 3.

   b  Alternate  tuning  criteria  may  be  used  (e.g.,   CLP,  Method  525,  or
      manufacturers'  instructions),  provided  that  method performance  is not
      adversely affected.
                                   TABLE 4.
                          CALIBRATION  CHECK COMPOUNDS
Base/Neutral Fraction
                  Acid Fraction
Acenaphthene
1,4-Dichlorobenzene
Hexachlorobutadiene
N-Nitrosodiphenylamine
Di-n-octyl phthalate
Fluoranthene
Benzo(a)pyrene
                  4-Chloro-3-methyl phenol
                  2,4-Dichlorophenol
                  2-Nitrophenol
                  Phenol
                  Pentachlorophenol
                  2,4,6-Trichlorophenol
                                  8270B - 37
                                            Revision 2
                                        September 1994

-------
                                   TABLE 5.
          SEMIVOLATILE  INTERNAL  STANDARDS WITH CORRESPONDING ANALYTES
                           ASSIGNED  FOR QUANTITATION
l,4-Dichlorobenzene-d4    Naphthalene-dE
                                Acenaphthene-d10
Aniline
Benzyl alcohol
Bis(2-chloroethyl) ether
Bis(2-chloroisopropyl)
                ether
2-Chlorophenol
1,3-Dichlorobenzene
1,4-Dichlorobenzene
1,2-Dichlorobenzene
Ethyl methanesulfonate
2-Fluorophenol (surr.)
Hexachloroethane
Methyl methanesulfonate
2-Methylphenol
4-Methylphenol
N-Nitrosodimethyl amine
N-Nitroso-di-n-propyl-
               amine
Phenol
Phenol-d6 (surr.)
2-Picoline
Acetophenone
Benzoic acid
Bis(2-chloroethoxy)methane
4-Chloroaniline
4-Chioro-3-methyl phenol
2,4-Dichlorophenol
2,6-Dichlorophenol
a,a-Dimethyl -
      phenethylamine
2,4-Dimethylphenol
Hexachlorobutadiene
Isophorone
2-Methyl naphthalene
Naphthalene
Nitrobenzene
Nitrobenzene-d8 (surr.)
2-Nitrophenol
N-Nitrosodibutyl amine
N-Nitrosopiperi dine
1,2,4-Trichlorobenzene
Acenaphthene
Acenaphthylene
1-Chloronaphthalene
2-Chloronaphthalene
4-Chlorophenyl
  phenyl ether
Dibenzofuran
Diethyl phthalate
Dimethyl phthalate
2,4-Dinitrophenol
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Fluorene
2-Fluorobiphenyl
        (surr.)
Hexachlorocyclo-
      pentadiene
1-Naphthylamine
2-Naphthylamine
2-Nitroaniline
3-Nitroaniline
4-Nitroaniline
4-Nitrophenol
Pentachlorobenzene
1,2,4,5-Tetra-
   chlorobenzene
2,3,4,6-Tetra-
   chlorophenol
2,4,6-Tribromo-
   phenol  (surr.)
2,4,6-Trichloro-
   phenol
2,4,5-Trichloro-
   phenol
(surr.) = surrogate
                                  8270B - 38
                                           Revision  2
                                       September  1994

-------
                                   TABLE  5.
                                  (Continued)
Phenanthrene-d
              10
Chrysene-d12
Perylene-d
                                                                    12
4-Aminobiphenyl
Anthracene
4-Bromophenyl phenyl
                ether
Di-n-butyl phthalate
4,6-Dinitro-2-methyl-
                phenol
Diphenylamine
Fluoranthene
Hexachlorobenzene
N-Nitrosodiphenylamine
Pentachlorophenol
Pentachloroni trobenzene
Phenacetin
Phenanthrene
Pronamide
Benzidine
Benzo(a)anthracene
Bis(2-ethylhexyl)
       phthalate
Butyl benzyl phthalate
Chrysene
3,3'-Dichlorobenzidine
p-Dimethylaminoazobenzene
Pyrene
Terphenyl-d14  (surr.)
Benzo(b)fluor-
    anthene
Benzo(k)fluor-
    anthene
Benzo(g,h,i)-
    perylene
Benzo(a)pyrene
Dibenz(a,j)acridine
Dibenz(a,h)-
     anthracene
7,12-Dimethylbenz-
    (a)anthracene
Di-n-octyl phthalate
Indeno(l,2,3-cd)
     pyrene
3-Methylchol-
     anthrene
(surr.) = surrogate
                                  8270B - 39
                                           Revision  2
                                       September  1994

-------
       TABLE 6.
QC ACCEPTANCE CRITERIA8
Compound
Acenaphthene
Acenaphthylene
Aldrin
Anthracene
Benz(a)anthracene
Benzo(b)fluoranthene
Benzo(k)fl uoranthene
Benzo(a)pyrene
Benzo(ghi)perylene
Benzyl butyl phthalate
/3-BHC
5-BHC
Bis(2-chloroethyl ) ether
Bis(2-chloroethoxy)methane
Bis(2-chloroisopropyl) ether
Bis(2-ethylhexyl) phthalate
4-Bromophenyl phenyl ether
2-Chloronaphthalene
4-Chlorophenyl phenyl ether
Chrysene
4,4'-DDD
4,4'-DDE
4,4'-DDT
Dibenzo( a, h) anthracene
Di-n-butyl phthalate
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1 ,4-Dichlorobenzene
3,3'-Dichlorobenzidine
Dieldrin
Diethyl phthalate
Dimethyl phthalate
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Di-n-octyl phthalate
Endosulfan sulfate
Endrin aldehyde
Fl uoranthene
Fluorene
Heptachlor
Heptachlor epoxide
Hexachlorobenzene
Hexachlorobutadiene
Test
cone.
(M9/L)
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
Limit
for s
(M9/L)
27.6
40.2
39.0
32.0
27.6
38.8
32.3
39.0
58.9
23.4
31.5
21.6
55.0
34.5
46.3
41.1
23.0
13.0
33.4
48.3
31.0
32.0
61.6
70.0
16.7
30.9
41.7
32.1
71.4
30.7
26.5
23.2
21.8
29.6
31.4
16.7
32.5
32.8
20.7
37.2
54.7
24.9
26.3
Range
for x
(M9/L)
60.1-132.3
53.5-126.0
7.2-152.2
43.4-118.0
41.8-133.0
42.0-140.4
25.2-145.7
31.7-148.0
D-195.0
D-139.9
41.5-130.6
D-100.0
42.9-126.0
49.2-164.7
62.8-138.6
28.9-136.8
64.9-114.4
64.5-113.5
38.4-144.7
44.1-139.9
D-134.5
19.2-119.7
D-170.6
D-199.7
8.4-111.0
48.6-112.0
16.7-153.9
37.3-105.7
8.2-212.5
44.3-119.3
D-100.0
D-100.0
47.5-126.9
68.1-136.7
18.6-131.8
D-103.5
D-188.8
42.9-121.3
71.6-108.4
D-172.2
70.9-109.4
7.8-141.5
37.8-102.2
Range
P, Ps
(%)
47-145
33-145
D-166
27-133
33-143
24-159
11-162
17-163
D-219
D-152
24-149
D-110
12-158
33-184
36-166
8-158
53-127
60-118
25-158
17-168
D-145
4-136
D-203
D-227
1-118
32-129
D-172
20-124
D-262
29-136
D-114
D-112
39-139
50-158
4-146
D-107
D-209
26-137
59-121
D-192
26.155
D-152
24-116
      8270B  - 40
    Revision 2
September 1994

-------




Compound
Hexachloroethane
Indeno(l,2,3-cd)pyrene
Isophorone
Naphthalene
Nitrobenzene
N-Ni trosodi -n-propyl amine
PCB-1260
Phenanthrene
Pyrene
1,2,4-Trichlorobenzene
4-Chloro -3 -methyl phenol
2-Chlorophenol
2,4-Chlorophenol
2,4-Dimethylphenol
2,4-Dinitrophenol
2-Methyl -4,6-dinitrophenol
2-Nitrophenol
4-NHrophenol
Pentachlorophenol
Phenol
2,4,6-Trichlorophenol
TABLE 6.
(Continued)
Test
cone.
(/*g/L)
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100


Limit
for s
(M9/L)
24.5
44.6
63.3
30.1
39.3
55.4
54.2
20.6
25.2
28.1
37.2
28.7
26.4
26.1
49.8
93.2
35.2
47.2
48.9
22.6
31.7


Range
for x
(M9/L)
55.2-100.0
D-150.9
46.6-180.2
35.6-119.6
54.3-157.6
13.6-197.9
19.3-121.0
65.2-108.7
69.6-100.0
57.3-129.2
40.8-127.9
36.2-120.4
52.5-121.7
41.8-109.0
D-172.9
53.0-100.0
45.0-166.7
13.0-106.5
38.1-151.8
16.6-100.0
52.4-129.2


Range
P» Ps
(%)
40-113
D-171
21-196
21-133
35-180
D-230
D-164
54-120
52-115
44-142
22-147
23-134
39-135
32-119
D-191
D-181
29-182
D-132
14-176
5-112
37-144
s     =     Standard deviation of four recovery measurements,  in

x     =     Average recovery  for four recovery measurements,  in ju9/L.

p, ps  =     Percent recovery  measured.

D     =     Detected; result  must be greater than  zero.

a     Criteria from 40  CFR  Part  136  for  Method  625.   These criteria are based
      directly on the method performance  data  in Table  7.  Where necessary, the
      limits for  recovery  have  been broadened to  assure  applicability of the
      limits to concentrations below those used to develop Table 7.
                                  8270B - 41
    Revision 2
September 1994

-------
                         TABLE 7.
METHOD ACCURACY AND PRECISION AS FUNCTIONS OF CONCENTRATION"
Compound
Acenaphthene
Acenaphthylene
Aldrin
Anthracene
Benz(a)anthracene
Chloroethane
Benzo(b)fluoranthene
Benzo(k)fluoranthene
Benzo(a)pyrene
Benzo(ghi)perylene
Benzyl butyl phthalate
0-BHC
5-BHC
Bis(2-chloroethyl) ether
Bis(2-chloroethoxy)methane
Bis(2-chloroisopropyl )
ether
Bis(2-ethylhexyl)
phthalate
4-Bromophenyl phenyl
ether
2-Chloronaphthalene
4-Chlorophenyl phenyl
ether
Chrysene
4,4'-DDD
4,4'-DDE
4,4'-DDT
Dibenzo( a, h) anthracene
Di-n-butyl phthalate
1,2-Dichlorobenzene
1 ,3-Dichlorobenzene
1,4-Dichlorobenzene
3,3'-Dichlorobenzidine
Dieldrin
Diethyl phthalate
Dimethyl phthalate
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Di-n-octyl phthalate
Endosulfan sulfate
Endrin aldehyde
Fluoranthene
Accuracy, as
recovery, x'
(M9/L)
0.96C+0.19
0.89C+0.74
0.78C+1.66
0.80C+0.68
0.88C-0.60
0.99C-1.53
0.93C-1.80
0.87C-1.56
0.90C-0.13
0.98C-0.86
0.66C-1.68
0.87C-0.94
0.29C-1.09
0.86C-1.54
1.12C-5.04
1.03C-2.31

0.84C-1.18

0.91C-1.34

0.89C+0.01
0.91C+0.53

0.93C-1.00
0.56C-0.40
0.70C-0.54
0.79C-3.28
0.88C+4.72
0.59C+0.71
0.80C+0.28
0.86C-0.70
0.73C-1.47
1.23C-12.65
0.82C-0.16
0.43C+1.00
0.20C+1.03
0.92C-4.81
1.06C-3.60
0.76C-0.79
0.39C+0.41
0.76C-3.86
0.81C+1.10
Single analyst
precision, s/
(M9A)
O.lSx-0.12
0.24x-1.06
0.27X-1.28
0.21X-0.32
O.lSx+0.93
0.14X-0.13
0.22X+0.43
0.19X+1.03
0.22X+0.48
0.29x4-2.40
0.18x4-0.94
0.20X-0.58
0.34X+0.86
0.35X-0.99
0.16X+1.34
0.24X+0.28

0.26X+0.73

0.13X+0.66

0.07X+0.52
0.20X-0.94

0.28X+0.13
0.29X-0.32
0.26X-1.17
0.42X+0.19
0.30X+8.51
0.13X+1.16
0.20X+0.47
0.25X+0.68
0.24X+0.23
0.28X+7.33
0.20X-0.16
0.28X+1.44
0.54X+0.19
0.12X+1.06
0.14X+1.26
0.21X+1.19
0.12X+2.47
0.18x+3.91
0.22X-0.73
Overall
precision,
S' (M9/L)
0.21X-0.67
0.26X-0.54
0.43X+1.13
0.27X-0.64
0.26X-0.21
0.17X-0.28
0.29X+0.96
0.35X+0.40
0.32X+1.35
0.51X-0.44
0.53X+0.92
0.30X+1.94
0.93X-0.17
0.35X+0.10
0.26X+2.01
0.25X+1.04

0.36X+0.67

0.16X+0.66

0.13X+0.34
0.30X-0.46

0.33X-0.09
0.66X-0.96
0.39X-1.04
0.65X-0.58
0.59X+0.25
0.39X+0.60
0.24X+0.39
0.41X+0.11
0.29X+0.36
0.47X+3.45
0.26X-0.07
0.52X+0.22
1.05X-0.92
0.21x4-1.50
0.19x4-0.35
0.37X+1.19
0.63X-1.03
0.73X-0.62
0.28X-0.60
                        8270B - 42
    Revision 2
September 1994

-------




Compound
Fluorene
Heptachlor
Heptachlor epoxide
Hexachlorobenzene
Hexachlorobutadiene
Hexachloroethane
Indeno(l,2,3-cd)pyrene
Isophorone
Naphthalene
Nitrobenzene
N-Nitrosodi-n-propylamine
PCB-1260
Phenanthrene
Pyrene
1,2,4-Trichlorobenzene
4- Chi oro -3 -methyl phenol
2-Chlorophenol
2,4-Dichlorophenol
2,4-Dimethylphenol
2,4-Dinitrophenol
2-Methyl-4,6-dinitrophenol
2-Nitrophenol
4-Nitrophenol
Pentachlorophenol
Phenol
2,4,6-Trichlorophenol
TABLE 7.
(Continued)
Accuracy, as
recovery, x'
(M9/U
0.90C-0.00
0.87C-2.97
0.92C-1.87
0.74C+0.66
0.71C-1.01
0.73C-0.83
0.78C-3.10
1.12C+1.41
0.76C+1.58
1.09C-3.05
1.12C-6.22
0.81C-10.86
0.87C+0.06
0.84C-0.16
0.94C-0.79
0.84C+0.35
0.78C+0.29
0.87C-0.13
0.71C+4.41
0.81C-18.04
1.04C-28.04
0.07C-1.15
0.61C-1.22
0.93C+1.99
0.43C+1.26
0.91C-0.18


Single analyst
precision, sr'
(M9/L)
0.12x+0.26
0.24x-0.56
0.33X-0.46
0.18x-0.10
0.19x+0.92
0.17X+0.67
0.29X+1.46
0.27x+0.77
0.21x-0.41
0.19x+0.92
0.27x+0.68
0.35x4-3.61
0.12X+0.57
0.16x+0.06
0.15x+0.85
0.23x+0.75
0.18X+1.46
O.lSx+1.25
0.16X+1.21
0.38X+2.36
O.lOx+42.29
0.16X+1.94
0.38X+2.57
0.24x+3.03
0.26X+0.73
0.16x+2.22


Overall
precision,
S' (M9/L)
0.13X+0.61
O.SOx-0.23
0.28X+0.64
0.43X-0.52
0.26X+0.49
0.17X+0.80
O.SOx-0.44
0.33X+0.26
0.30X-0.68
0.27x+0.21
0.44x+0.47
0.43X+1.82
0.15X+0.25
0.15X+0.31
0.21X+0.39
0.29X+1.31
0.28x+0,97
0.21X+1.28
0.22X+1.31
0.42x4-26.29
0.26X+23.10
0.27X+2.60
0.44X+3.24
0.30X+4.33
0.35X+0.58
0.22X+1.81
X'
S'


C

x
Expected  recovery  for  one  or  more  measurements  of  a  sample
containing a concentration  of C, in fJ.g/1.

Expected single  analyst  s_tandard deviation of measurements  at an
average concentration of x, in /j.g/1.

Expected interlaboratory standard deviation of measurements  at an
average concentration found of x, in
True value for the concentration, in iJ.g/1.

Average recovery  found for measurements  of samples containing  a
concentration of C, in M9/L.
                                  8270B - 43
                                                        Revision 2
                                                    September 1994

-------
                                   TABLE 8.
      SURROGATE  SPIKE  RECOVERY  LIMITS  FOR WATER AND SOIL/SEDIMENT SAMPLES

Surrogate Compound
Nitrobenzene-d5
2-Fluorobiphenyl
Terphenyl-du
Phenol -d6
2-Fluorophenol
2,4,6-Tribromophenol
Low/High
Water
35-114
43-116
33-141
10-94
21-100
10-123
Low/High
Soil/Sediment
23-120
30-115
18-137
24-113
25-121
19-122
                                   TABLE 9.
              EXTRACTION  EFFICIENCY  AND  AQUEOUS  STABILITY  RESULTS
COMPOUND
PERCENT RECOVERY
   ON DAY 0
AVG.      RSD
PERCENT RECOVERY
  ON DAY 7
AVG.    RSD
3-Amino-9-ethylcarbazo1e                  80
4-Ch1oro-l,2-phenylenediamine             91
4-Chloro-l,3-phenylenediamine             84
l,2-Dibromo-3-chloropropane               97
2-sec-Butyl-4,6-dinitrophenol             99
Ethyl parathion                          100
4,4'-Methylenebis(N,N-dimethyl aniline)   108
2-Methyl-5-nitroaniline                   99
2-Methylpyridine                          80
Tetraethyl dithiopyrophosphate            92
            8
            1
            3
            2
            3
            2
            4
           10
            4
            7
 73
108
 70
 98
 97
103
 90
 93
 83
 70
3
4
3
5
6
4
4
4
4
1
Data from Reference 8.
                                  8270B - 44
                           Revision 2
                       September  1994

-------
                              TABLE  10.
AVERAGE PERCENT RECOVERIES AND PERCENT RSDs FOR THE TARGET COMPOUNDS
  FROM  SPIKED CLAY  SOIL  AND TOPSOIL  BY  AUTOMATED SOXHLET EXTRACTION
                     WITH  HEXANE-ACETONE  (1:1)'
                                   Clay Soil
 Topsoil
Compound name
1,3-Dichlorobenzene
1,2-Dichlorobenzene
Nitrobenzene
Benzal chloride
Benzotrichloride
4-Chloro-2-nitrotoluene
Hexachl orocycl opentadi ene
2,4-Dichloronitrobenzene
3,4-Dichloronitrobenzene
Pentachl orobenzene
2,3,4,5-Tetrachloronitrobenzene
Benefin
alpha-BHC
Hexachl orobenzene
delta-BHC
Heptachlor
Aldrin
Isopropal in
Heptachlor epoxide
trans-Chlordane
Endosulfan I
Dieldrin
2,5-Dichlorophenyl-
4-nitrophenyl ether
Endrin
Endosulfan II
p,p'-DDT
2,3,6-Trichlorophenyl-
4'-nitrophenyl ether
2,3,4-Trichlorophenyl-
4'-nitrophenyl ether
Mi rex
Average
percent
recovery
0
0
0
0
0
0
4.1
35.2
34.9
13.7
55.9
62.6
58.2
26.9
95.8
46.9
97.7
102
90.4
90.1
96.3
129
110

102
104
134
110

112

104
Percent
RSD
_.
--
--
--


15
7.6
15
7.3
6.7
4.8
7.3
13
4.6
9.2
12
4.3
4.4
4.5
4.4
4.7
4.1

4.5
4.1
2.1
4.8

4.4

5.3
Average
percent
recovery
0
0
0
0
0
0
7.8
21.2
20.4
14.8
50.4
62.7
54.8
25.1
99.2
49.1
102
105
93.6
95.0
101
104
112

106
105
111
110

112

108
Percent
RSD
..
--
--



23
15
11
13
6.0
2.9
4.8
5.7
1.3
6.3
7.4
2.3
2.4
2.3
2.2
1.9
2.1

3.7
0.4
2.0
2.8

3.3

2.2
 The  operating  conditions  for  the  Soxtec  apparatus  were  as  follows:
 immersion time 45 min; extraction time 45 min; the sample size was 10 g;
 the spiking concentration was 500 ng/g, except for the surrogate compounds
 at 1000 ng/g,  compounds  23, 27,  and  28 at 1500 ng/g,  compound 3 at 2000
 ng/g, and compounds 1 and 2 at 5000  ng/g.
                             8270B - 45
    Revision 2
September 1994

-------
                            TABLE  11.
SINGLE LABORATORY ACCURACY AND PRECISION DATA FOR THE EXTRACTION
          OF SEMIVOLATILE ORGANICS FROM SPIKED CLAY BY
                METHOD 3541 (AUTOMATED SOXHLET)8
Compound name
Phenol
Bis(2-chloroethyl)ether
2-Chlorophenol
Benzyl alcohol
2-Methyl phenol
Bis(2-chloroisopropyl )ether
4-Methyl phenol
N-Nitroso-di-n-propylamine
Nitrobenzene
Isophorone
2-Nitrophenol
2,4-Dimethylphenol
Benzole acid
Bis(2-chloroethoxy) methane
2,4-Dichlorophenol
1,2,4-Trichlorobenzene
Naphthalene
4-Chloroaniline
4-Chloro -3 -methyl phenol
2 -Methyl naphthalene
Hexachl orocycl opentad i ene
2,4,6-Trichlorophenol
2,4,5-Trichlorophenol
2-Chloronaphthalene
2-Nitroaniline
Dimethyl phthalate
Acenaphthylene
3-Nitroaniline
Acenaphthene
2,4-Dinitrophenol
4-Nitrophenol
Dibenzofuran
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Diethyl phthalate
4-Chlorophenyl-phenyl ether
Fluorene
4-Nitroaniline
4, 6-Dinitro-2-methyl phenol
N-Nitrosodiphenylamine
4-Bromophenyl-phenyl ether
Average
percent
recovery
47.8
25.4
42.7
55.9
17.6
15.0
23.4
41.4
28.2
56.1
36.0
50.1
40.6
44.1
55.6
18.1
26.2
55.7
65.1
47.0
19.3
70.2
26.8
61.2
73.8
74.6
71.6
77.6
79.2
91.9
62.9
82.1
84.2
68.3
74.9
67.2
82.1
79.0
63.4
77.0
62.4
Percent
RSD
5.6
13
4.3
7.2
6.6
15
6.7
6.2
7.7
4.2
6.5
5.7
7.7
3.0
4.6
31
15
12
5.1
8.6
19
6.3
2.9
6.0
6.0
5.2
5.7
5.3
4.0
8.9
16
5.9
5.4
5.8
5.4
3.2
3.4
7.9
6.8
3.4
3.0
                           8270B - 46
    Revision 2
September 1994

-------
                             Table  11.  (Continued)
Compound name
Average
percent
recovery
             Percent
               RSD
Hexachlorobenzene
Pentachlorophenol
Phenanthrene
Anthracene
Di-n-butyl phthalate
Fluoranthene
Pyrene
Butyl benzyl phthalate
3,3'-Dichlorobenzidine
Benzo(a)anthracene
Bis(2-ethylhexyl) phthalate
Chrysene
Di-n-octyl phthalate
Benzo(b)fluoranthene
Benzo(k)fluoranthene
Benzo(a)pyrene
Indeno(l,2,3-cd)pyrene
Dibenzo(a,h)anthracene
Benzo(g,h,i)perylene
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
Hexachloroethane
Hexachlorobutadiene
  72.6
  62.7
  83,
  96.
  78.
  87.
 102
  66.
  25.
  73.4
  77.2
  76.2
  83.1
.9
.3
.3
.7

.3
.2
  82.7
  71.7
  71.7
  72.2
  66.7
  63.9
   0
   0
   0
   0
   0
 3.7
 6.1
 5.4
 3.9
40
 6.9
 0.8
 5.2
11
 3.8
 4.8
 4.4
 4.8
 5.0
 4.1
 4.1
 4.3
 6.3
 8.0
a     Number  of  determinations was three.   The operating  conditions  for the
      Soxtec apparatus were as follows:   immersion time 45 min; extraction time
      45 min; the  sample  size  was  10  g  clay soil;  the spike concentration was
      6 mg/kg per compound.  The sample  was  allowed to equilibrate 1 hour after
      spiking.

Data taken from Reference 9.
                                  8270B - 47
                   Revision 2
               September  1994

-------
                                     FIGURE 1.
       GAS CHROMATOGRAM OF  BASE/NEUTRAL AND ACID  CALIBRATION STANDARD
R1C
        8:26:09
     :  BASE ACID STD. 2UL/!oNC- UL
CWCS.:
R*€E: C  1,2768  LflfcEU N  6. 4.8
CUT*: 5l&H*C«w786 kl
CHL!: 51BHS6S878S 13
             SCANS 2ue TO 2786
0.  1.6 J 6  E»c£: U 28.  3
                                    8270B  -  48
                                    Revision 2
                                September 1994

-------
                                  METHOD 82708
SEMIVOLATILE ORGANIC COMPOUNDS BY GAS CHROMATOGRAPHY/MASS SPECTROMETRY
                     (GC/MS):   CAPILLARY  COLUMN  TECHNIQUE
    7.1 Prepare sample
    using Method 3540,
      3541, or 3550.
7.1 Prepare sample
using Method 3510
    or 3520.
                                  7.1 Prepare sample
                                  using Method 3540,
                                 3541,  3650, or 3580.
                                     7.2 Cleanup
                                      extract.
                                   7.3 Set GC/MS
                                 operating conditions;
                                    perform initial
                                     calibration.
                                   7.4 Perform daily
                                 calibration with SPCCs
                                   and CCCs prior to
                                  analysis of samples.
                                   8270B - 49
            Revision 2
       September 1994

-------
                                   METHOD 82708
                                    (Continued)
                            7.5.1 Screen extract
                           on GC/FID or GC/PID to
                           eliminate samples that
                            are too concentrated.
                            7.5.3 Analyze extract
                              by GC/MS, using
                            appropriate fused-silica
                              capillary column.
7.5.4 Dilute
  Extract.
    7.5.4
Does response
 exceed initial
  calibration
   curve?
7.6.1 Identify
analyte by comparing
the sample and standard
mass spectra.
^
r
7.6.2 Calculate
concentration of each
individual analyte;
report results.
>
f
                                (   Stop    J
                                     8270B -  50
                                                Revision  2
                                           September  1994

-------