THE ANALYSIS OF TRIHALOMETHANES IN FINISHED
WATERS BY THE PURGE AND TRAP METHOD
^ PRO^
U. S. ENVIRONMENTAL PROTECTION AGENCY
ENVIRONMENTAL MONITORING AND SUPPORT LABORATORY
CINCINNATI, OHIO 4526S

-------
Foreword
This method has been prepared by the staff of the Environmental
Monitoring and Support Laboratory-Cincinnati at the request of the Office
of Water Supply. Comments and suggestions offered by the Municipal
Environmental Research Laboratory and the Division of Technical Support
on the August 1, 1977 draft are gratefully acknowledged.
The procedure represents the current state-of-the-art, but as time
progresses improvements are anticipated. Users are encouraged to identify
problems and assist in updating the method by contacting the Environmental
Protection Agency, Environmental Monitoring and Support Laboratory,
Cincinnati, Ohio 45268.

-------
The Analysis of Trihalomethanes in Drinking
Water by the Purge and Trap Method
Scope
1.1	This method is applicable in the determination of chloroform,
dichlorobromomethane, dibromochlororaethane, and bromoform in
drinking water or raw source water. The concentration of these
four compounds may be totaled to determine total trihalomethanes
(TTHM).
1.2	Though the actual detection limits are highly dependent upon
the gas chromatographic column and detector employed, the method
can be used over a concentration range of approximately 0.1 to
1500 micrograms per liter.
1.3	Well in excess of 100 different water supplies have been analyzed
using this method. Supplementary analyses using gas chromatog-
raphy mass spectrometry (GC/MS) have shown that there is no
evidence of interference in the determination of trihalomethanes.
For this reason it is not necessary to analyze the raw source
water as is required with the interim Liquid/Liquid Extraction
Method.
Summary
2.1 An extraction/concentration technique is incorporated within
the method which enhances the amounts of trihalomethanes injected
into the gas chromatograph by a factor of 1000 over direct injec-
tion gas chromatography and by a factor of 200 over the interim
Liquid/Liquid Extraction Method.

-------
2.2	Trihaloraethanes are extracted by an inert gas which is bubbled
through the aqueous sample. The trihaloraethanes, along with
other organic constituents which exhibit low water solubility
and a vapor pressure significantly greater than water, are
efficiently transferred from the aqueous phase to the gaseous
phase. These compounds are swept from the purging device and
are trapped in a short column containing a suitable sorbent.
After a predetermined period of time, the trapped components
are thermally desorbed and backflushed onto the head of a gas
chromatographic column and separated under programmed conditions.
Measurement is accomplished with a halogen specific detector
such as electrolytic conductivity or microcoulometric titration.
2.3	Confirmatory analyses are performed using dissimilar columns,
or by mass spectrometry.'.
2.4	Aqueous standards and unknowns are extracted and analyzed under
identical conditions in order to compensate for extraction losses.
2.5	The total analysis time is approximately 35 minutes per sample.
Interferences
3.1 Impurities contained in the purge gas and organic compounds
out-gasing from the plumbing ahead of the trap usually account
for the majority of contamination problems. The presence of
such interferences are easily monitored as a part of the quality
control program. Sample blanks are normally run between each
-set of samples. When a positive trihalomethane response is
noted in the sample blank, the analyst should analyze a method

-------
blank. Method blanks are run by charging the purging device
with organic-free water and analyzing in the normal manner.
Whenever trihalomethanes are noted in the method blank,
the analyst should change the purge gas source and regenerate
the molecular sieve purge gas filter. Subtracting the blank
values is not recommended. The use of non-TFE plastic tubing,
non-TFE thread sealants, or flow controllers with rubber com-
ponents should be avoided since such materials generally out-gas
organic compounds which will be concentrated in the trap during
the purge operation. Such out-gasing problems are common when-
ever new equipment is put into service; as time progresses minor
out-gasing problems generally cure themselves.
3.2	Several instances of accidental sample contamination have been
noted and attributed to diffusion of volatile organics through
the septum seal and into the sample during shipment and storage.
The sample blank is used as a monitor for this problem.
3.3	. For compounds that are not efficiently purged, such as broraoform,
small variations in sample volume, purge time, purge flow rate,
or purge temperature can affect the analytical result. Therefore,
samples and standards must be analyzed under identical conditions.
3.4	Cross contamination can occur whenever high level and low level
samples are sequentially analyzed. To reduce the likelihood
of this, the purging device and sample syTinge should be rinsed
out twice between samples with organic free water. Whenever an
unusually concentrated sample is encountered, it is highly

-------
- 4 -
recommended that it be followed by a sample blank analysis to
insure that sample cross contamination does not occur. For
samples containing large amounts of water soluble materials,
it may be necessary to wash out the purging device with a
soap solution, rinse with distilled water, and then dry in a
105°C oven between analyses.
3.S Qualitative misidentifications are a problem in using gas
chromatographic analysis. Whenever samples whose qualitative
nature is unknown are analyzed, the following precautionary
measures should be incorporated into the analysis.
3.5.1	Perform duplicate analyses using the two recommended
columns (4.2.1 and 4.2.2) which provide different
retention order and retention times for the tri-
halomethanes and other organohalides.
3.5.2	Use element specific detectors.
3.5.3	Whenever possible use GC/MS techniques which provide
unequivocal qualitative identification^.
4. Apparatus
4.1 The purge and trap equipment consists of three separate pieces
of apparatus: the purging device, trap, and desorber. Con-
struction details for a purging device and an easily automated
trap-desorber hybrid which has proven to be exceptionally
efficient and reproducible are shown in Figures 1 through 4
and described in 4.1.1 through 4.1.3. An earlier acceptable
version of the above mentioned equipment is described in Ref. 1.

-------
- 5 -
4.1.1 Purging Device
Construction details are given in Figure 1 for an
all-glass 5 ml purging device. The glass frit installed
at the base of the sample changer allows finely divided
gas bubbles to pass through the sample while the sample
is restrained above the frit. Gaseous volumes above the
sample are kept to a minimum to eliminate dead volume
effects, yet allowing sufficient space for most foams to
disperse. The inlet and exit ports are constructed from
heavy walled 1/4 inch glass tubing so that leak-free
removable connections can be made using "finger-tight"
compression fittings containing Teflon ferrules. The
removable foam trap is used to control samples
that foam.
' 4.1.2 Trapping Device
The trap (Figure 2) is a short gas chromatographic
column which at 22°C retards the flow of the compounds
of interest while venting the purge gas and, depending on
which sorbent is used, much of the water vapor. The
tTap should be constructed with a low thermal mass so
that it can be rapidly heated for efficient desorption,
then rapidly cooled to room temperature for recycling.
Variations in the trap ID, wall thickness, sorbents,
soTbent packing order, and sorbent mass could adversely

-------
- 6 -
affect the trapping and desorption efficiencies for
compounds discussed in this text. For this reason, it
is important to faithfully reproduce the trap configurations
recommended in Figure 2. The general purpose trap con-
taining only Tenax is used for the trihalomethanes and
compounds that boil above 30 °C. Compounds that boil
below 30°C are not strongly sorbed by Tenax; therefore,
such compounds pass through Tenax traps and are vented
under normal purging conditions. If such compounds are
to be determined, in addition to the trihalomethanes,
the multi-purpose trap should be used. Grade-15 silica
gel effectively retards the flow of most gaseous substituted
organics at 22°C while allowing efficient desorption at 180"C.
Higher boiling compounds do not efficiently desorb from the
silica gel. The Tenax-silica gel multi-purpose trap
utilizes the sorptive properties of the two sorbents
providing a trap which effectively sorbs and desorbs a
wide variety of organic compounds.
4.1.3 Desorber assembly - Details for the desorber are shown in
Figures 2, 3, and 4. With the 6-port valve in the Purge-
Sorb position (Figure 3), the effluent from the purging
device passes through the trap where the flow rate of
the organics is retarded. The GG carrier gas also passes
through the 6-port valve and is returned to the GC. With

-------
- 7 -
the 6-port valve in the Purge-Sorb position, the operation
of the GC is in no way impaired; therefore, routine liquid
injection analyses can be performed using the gas
chromatograph. After the sample has been purged, the
6-port valve is turned to the desorb position (Figure 4).
In this configuration, the trap is coupled in series
with the gas chromatographic column allowing the
carrier gas to back-flush the trapped materials into the
analytical column. Just as the valve is actuated the
power is turned on to resistance wire wrapped around the
trap. The power is supplied by an electronic temperature
controller. Using this device, the trap is rapidly heated
to 180*0 and then maintained at 180°C with minimal temper-
ature overshoot. The trapped compounds are released as a
"plug" to the gas chromatograph by this heat and backflush
step. Normally, packed columns with theoretical efficiencies
near 500 plates/foot under programmed temperature conditions
can accept such desorb injections without altering peak
geometry. Substituting a non-controlled power supply,
.such as a manually-operated variable transformer, will
provide non-reproducible retention times and poor quanti-
tative data unless Injection Procedure (8.9.2) is used.
-4.1.4 A commercial device manufactured by Tekmar has been tested
and shown to be equivalent: Tekmar, P.O. Box 37202,
Cincinnati, Ohio 45202. This device or its equivalent
may be used.

-------
- 8 -
4.2. Gas chromatograph - The chromatograph must be temperature
programmable and equipped with a halide specific detector.
4.2.1 Column 1 is an unusually efficient column which provides
outstanding separations for a wide variety of organic
compounds. Because of its ability to resolve tri-
halomethanes from other organochlorine compounds, column
I should be used as the primary analytical column (see
Figure 5).
4.2.1.1 Column I parameters: Dimensions - eight feet
long x 0.1 inch ID stainless steel or glass
tubing. Packing - 0.2% Carbowax 1500 on
Carbopack-C (80/100) mesh. Carrier Gas -
helium at 40 ml/minute. Temperature program
sequence - 60°C isothermal for 3 minutes, pro-
gram at 8°C/minute to 160°C, then hold for
2 minutes or until all compounds have eluted.
NOTE: It has been found that during handling,
packing, and programming, active sites are
exposed on the.Carbopack-C packing. This
results in poor resolution of constituents
and poor peak geometry. To correct this,
place a 1 ft. 0.125 in. 00 x 0.1 in. ID stainless
steel column packed with 3% Carbowax 1500 on
ChromosoTb-W 60/80 mesh in series before the
CaTbopack-C column. Condition the precolumn

-------
- 9 -
and the Carbopack columns with carrier gas
flow at 190°C overnight. The two columns may
be retained in series for routine analyses.
Trihalomethane retention times are listed in
Table 1.
4.2.2 Column II provides unique organohalide-trihalomethane
separations when compared to those obtained from Column I
(see Figure 6). However, since the resolution between
various compounds is generally not as good as those with
Column I, it is recommended that column II be used as a
qualitative confirmatory column for unknown samples when
GC/MS confirmation is not possible.
4.2.2.1 Column II parameters: Dimensions - six feet
long x 0.1 inch ID stainless steel or glass.
Packing - n-octane on Porisil-C (100/120 mesh).
Carrier Gas - helium at 40 cc/minute. Tempera-
ture program sequence - S0°C isothermal for
3 minutes, program at 6'/minutes to 170°C,
then hold for 4 minutes or until all compounds
have eluted. Trihalomethane retention times
are listed in Table 1.
4.3 Sampling containers - 40 ml scTew cap vials sealed with Teflon
faced silicone septa.
Vials and caps - Pierce #13075 or equivalent
Septa - Pierce #12722 or equivalent

-------
- 10 -
4.4	Syringes - 5-ral hypodermic with luerlok tip.(2 each).
4.5	Micro syringes - 10, 25, 100 yl.
4.6	2-way syringe valve with luer ends (3 each).
5. Reagents and Materials
5.1	Porous polymer packing 60/80 mesh chromatographic grade Tenax GC
(2,6-diphenylene oxide).
5.2	Three percent OV-1 Chromosorb-W 60/80 mesh.
5.3	0.2% Carbowax 1500 on Carbopack-C (80/100 mesh) available from
Supelco; request Batch #R-1579.
5.4	N-octane on Porasil-C (100/120 mesh) available from Waters
Associates.
5.5	Three percent Carbowax 1500 on Chromosorb-W (60/80 mesh).
5.6	Dechlorinating compound-crystalline sodium thiosulfate, A.C.S.
Reagent Grade.
5.7	Activated carbon - Filtrasorb-200, available from Calgon Corp.,
Pittsburgh, PA, or equivalent.
5.8	Organic-free water
5.8.1	OTganic-free water is generated by passing tap water
through a carbon filter bed containing about 1 lb. of
activated carbon.
5.8.2	A Millipore Super-Q Water System or its equivalent may
be used to generate organic free deionized water.
NOTE: Test organic free water daily by analyzing
according to this method. See (8).

-------
- 11 -
5.9	Standards
5.9.1	Bromoform - A.C.S. Reagent Grade
5.9.2	Bromodichloromethane 97% - available from Aldrich
Chemical Company.
5.9.3	Chiorodibromomethane - available from Columbia Chemical
Inc., Columbia, SC.
5.9.4	Chloroform - A.C.S. Reagent Grade.
5.10	Standard Stock Solutions
5.10.1	Place about 9.8 ml of methyl alcohol into a ground glass
stoppered 10 ml volumetric flask.
5.10.2	Allow the flask to stand unstoppered about 10 minutes
or until all alcohol wetted surfaces have dried.
5.10.3	Weigh the flask to the nearest 0.1 rag.
5.10.4	Using a 100 yl syringe, immediately add 2 drops of the
reference standard to the flask, then reweigh. Be sure
that the 2 drops fall directly into the alcohol without
contacting the neck of the flask.
5.10.5	Dilute to volume, stopper, then mix by inverting the
flask several times.
NOTE: Because of the toxicity of trihalomethanes, it
is necessary to prepare primary dilutions in a hood.
It is further recommended that a NI0SH/MESA approved
toxic gas respirator be used when the analyst handles
high concentrations of such materials:

-------
- 12 -
5.10.6	Calculate the concentration in micrograms per microliter
from the net gain in weight.
5.10.7	Store the solution at 4°C.
NOTE: , All standard solutions prepared in methyl alcohol
are stable up to 4 weeks when stored under these conditions.
They should be discarded after that time has elapsed.
5.11	Calibration Standards
5.11.1	Prepare, from the standard stock solutions, secondary
dilution mixtures in methyl alcohol so that a 20 pi
injection into, 100 ml of organic-free water will generate
a calibration standard which produces a response close
(±10%) to that of the unknowns.
NOTE: Aqueous standards are not stable and should be
discarded after one hour.
5.11.2	Purge and analyze the aqueous calibration standards in
the same manner as the unknowns.
5.12	Quality Check Standard (2.0 ug/1)
5.12.1	From the standard stock solutions, prepare a secondary
dilution in methyl alcohol containing 10 ng/yl of each
trihalomethane.
5.12.2	Daily, inject 20.0 ul of this mixture into 100.0 ml
of organic-free water and analyze according to the
Procedure (8). (See 5.11.1 note.)
6. Sample Collection and Handling
6.1 The sample containers should have a total volume in excess of 50 ml.

-------
- 13 -
6.1.1 Narrow mouth screw cap bottles with the TFE fluorocarbon
face silicone septa cap lineTS are strongly recommended.
Crimp-seal serum vials with TFE fluorocarbon faced septa
are acceptable if the seal is properly made and maintained
during shipment.
6.2	Sample Bottle Preparation
6.2.1	Wash all sample bottles and TFE seals in detergent.
Rinse with tap water and finally with distilled water.
6.2.2	Allow the bottles and seals to air dry at room temperature,
then place in a 105°C oven for one hour, then allow to
cool in an area known to be free of organics.
NOTE: Do not heat the TFE seals for extended periods
of time (>1 hour) because the silicone layer slowly
degrades at 105°C.
6.2.3	When cool, seal the bottles using the TFE seals that
will be used for sealing the samples.
6.3	Sample Preservation - Sodium thiosulfate, a chemical dechlorinating
agent, is added to the sample in order to arTest the formation
of trihalomethanes after sample collection (Ref. 2). If
chemical preservation is employed, the preservative is also
added to the blanks. The chemical preservative (2.5 to 3 rag/40 ml)
is. added to the empty sample bottles just prior to shipping to
the sampling site. Do not add sodium thiosulfate to samples
when data on maximum trihalomethane formation is desired.

-------
- 14 -
Sample Collection
6.4.1	Collect all samples in duplicate.
6.4.2	Fill the sample bottles in such a manner that no air
bubbles pass through the sample as the bottle is filled.
6.4.3	Seal the bottles so that no air bubbles are entrapped
in it.
6.4.4	Maintain the hermetic seal on the sample bottle until
analysis.
6.4.5	Sampling from a water tap.
6.4.5.1 Turn on water and allow the system to flush.
When the temperature of the water has stabilized,
adjust the flow to about 500 ml/minute and
collect duplicate samples from the flowing stream.
6.4.6	Sampling from an open body of water.
6.4.6.1 Fill a 1-quart wide-mouth bottle with sample
from a representative area. Carefully fill
duplicate sample bottles from the 1-quart
bottle as noted in 6.4.2.
6.4.7	' If preservative has been added to the sample bottles,
fill with sample just to overflowing, seal the bottle,
and shake vigorously for 1 minute.
6.4.8	Sealing practice for septum seal screw cap bottles.
6.4.8.1 Open the bottle and fill to overflowing, place
on a level surface, position the TFE side of

-------
- IS -
the septum seal upon the convex sample meniscus
and seal the bottle by screwing the cap on
tightly.
6.4.8.2 Invert the sample and lightly tap the cap on a
solid surface. The absence of entrapped air
indicates a successful seal. If bubbles are
present, open the bottle, add a few additional
drops of sample and reseal bottle as above.
6.4.9 Blanks
6.4.9.1	Prepare blanks in duplicate at the laboratory
by filling and sealing sample bottles with
organic-free water just prior to shipping the
sample bottles to the sampling site.
6.4.9.2	If the sample is to be preserved, add an identical
amount of preservative to the blanks.
6.4.9.3	Ship the blanks to and from the sampling
site along with the sample bottles.
6.4.9.4	Store the blanks and the samples collected at a
given site (sample set), together. A sample set
is defined as all the samples collected at a
given site (i.e., at a water treatment plant,
the duplicate raw source waters, the duplicate
finished waters and the duplicate blank samples
comprise the sample set).

-------
- 16 -
Conditioning Traps'
7.1	Condition newly packed traps overnight at 200°C with an inert
gas flow of at least 20 ml/min.
7.1.1 Vent the trap effluent to the room, not to the analytical
column.
7.2	Prior to daily use, condition traps 10 minutes while backflushing
at 180°C.
7.2.1 The trap may be vented to the analytical column; however,
after conditioning the column must be programmed prior
to use.
Extraction and Analysis
8.1	Adjust the purge gas (nitrogen or helium) flow rate to 50 ml/min.
8.2	Attach the trap inlet to the purging device. Turn the valve to
the purge-sorb position (Figure 3).
8.3	Open the syringe valve located on the purging device sample
introduction needle.
8.4	Remove the plungers from two 5 ml syringes and attach a closed
syringe valve to each.
8.5	Open the sample bottle and carefully pour the sample into one
of the syTinge barrels until it overflows. Replace the syringe
plunger and compress the sample. Open the syringe valve and
vent any residual air while adjusting the sample volume to
5.0 ml. Close the valve.

-------
- 17 -
8.6	Fill the second syringe in an identical manner from the same
sample bottle. This second syringe is reserved for a duplicate
analysis, if necessary.
8.7	Attach the syringe-valve assembly to the syringe valve on the
purging device.
8.8	Open the syringe valve and inject the sample into the purging
chamber. Close both valves. Purge the sample for 11.0 ± .05
minutes.
8.9	After the 11 minute purge time, attach the trap to the chromato-
graph (turn the valve to the desorb position) and introduce the
trapped materials to the GC column by rapidly heating the trap
to 180°C while backflushing the trap with an inert gas between
20 and 60 ml/min for 4 minutes.
8.9.1	If the trap can be rapidly heated to 180°C and maintained
at this temperature, the GC analysis can begin as the
sample is desorbed, i.e., the column is at the initial
60®C operating temperature. The equipment described in
Figure 4 will perform accordingly.
8.9.2	With other types of equipment (see 4.1.4 and Reference 1)
where the trap is not rapidly heated or is not heated in
a reproducible manner, it is necessary to transfer the
contents of the trap into the analytical column at 30°C
where it is once again trapped. Once the transfer is
complete (4 minutes), the column is rapidly heated to
the initial operating temperature for analysis.

-------
- 18 -
8.9.3 If injection procedure 8.9.1 is used and the early
eluting peaks in the resulting chromatogram have poor
geometry or variable retention times, then method 8.9.2
should be used.
8.10	After the extracted sample is introduced into the gas chromatograph,
empty the gas purging device using the sample introduction syTinge,
follow by two 5 ml flushes of organic-free water. When the
purging device is emptied, leave the syTinge valve open allowing
the purge gas to vent through the sample introduction needle.
8.11	Analyze each sample and sample blank from the sample set in an
identical manner (see 6.4.9.4) on the same day.
8.12	Prepare calibration standards from the standard stock solutions
(5.10) in organic-free water that are close to the unknown in
trihalomethane composition and concentration (9.1). The concen-
trations should be such that only 20 yl or less of the secondary
dilution need be added to 100 ml of organic-free water to produce
a standard at the same level as the unknown.
9. Analytical Quality Control
9.1 Analyze the 2 yg/1 quality check sample daily before any samples
are analyzed. Instrument status checks and lower limit of
detection estimations based upon response factor calculations
at two times the signal to noise ratio are obtained from these
data. In addition, response factor data obtained from the 2 yg/1
quality check standard can be used to estimate the concentration

-------
- 19 -
of the unknowns. From this information the appropriate standard
dilutions can be determined.
9.2	Analyze the EMSL-Cincinnati volatile oTganics quality control
samples or their equivalent on a quarterly basis.
9.3	Analyze the sample blank to monitor for potential interferences
as described in section 3.1, 3.2 and 3.4.
Calculations
10.1	Quantify the unknowns by comparing the peak height of the
unknowns to the standard peak height (8.12). Round off the data
to the nearest yg/1.
* (cone. std. Mg/1)
10.2	Report the results obtained from the EMSL Quality Control Sample
and the lower limit of detection estimates along with the data
for the unknown samples.
10.3	Calculate the total trihalomethane concentration (TTHM) by
summing the 4 individual trihalomethane concentrations in yg/1.
TTHM (yg/1) = (Cone. CHCU) + (Cone. CHBrCl2) + (Cone. CHBr2Cl) +
Cone. CHBr^)•
10.4	TTHM (ffig/1) =
10.5	Calculate the limit of detection (L0D) for each trihalomethane
not detected vising the following criteria:
-2 fHHeffJ
where B = peak height (mm) of 2 yg/1 quality check standard
A = 2 times the noise level in (mm) at the exact retention

-------
- 20 -
time of the trihalomethane or the baseline displace-
ment in (mm) from the theoretical zero at the exact
retention time of the trihalomethane.
ATT = Attenuation factor
Accuracy and Precision
11.1 One liter of organic-free water was dosed with the trihalomethanes;
The dosed water was used to fill septum seal vials which were
stored under ambient conditions. The dosed samples were randomly-
analyzed over a 2-week period of time. The data listed in Table II
reflect the errors due to the analytical procedure and storage.

-------
- 21 -
Table 1
Retention Data (minutes)
Column I	Column II
Chloroform 8.2	12.2
Bromodichlororaethane 10.8	14.7
Dichlorobromomethane 13.2	16.6
Bromoform 15.7	19.2

-------
- 22 -
Table II
Accuracy and Precision for Trihalomethanes
Chloroform
Dose	Number*	Mean	Standard
yg/1	samples	yg/1	deviation
1.19	12	1.21	0.14
11.9	8	11.3	0.16
119	11	105	7.9
Bromodichloromethane
Dose	Number*	Mean	Standard
yg/1	samples	yg/1	deviation
1.60	12	1.52	0.05
16.0	8	15.1	0.39
160	11	145	10.2
Chlorodibromomethane
Dose	Number*	Mean	Standard
yg/1	samples	yg/1	deviation
1.96	12	1.91	0.09
19.6	8	19.1	0.70
196	11	185	10.6
Bromoform
Dose	Number*	Mean	Standard
yg/1	samples	yg/1	deviation
2.31	12	2.33	0.16
23.1	8	22.5	1.38
231	11	223	16.3
•Single laboratory data

-------
- 23 -
References
Bellar, T. A., Lichtenberg, J. J., Determining Volatile Organics
at the Microgram per Litre Levels by Gas Chromatography, Journal
AWWA., 66, 739 (December, 1974).
Identification and Analysis of Organic Pollutants in Water,
Keith, L. H., Ann Arbor Science, p. 87 (1976).

-------
COLUMN: n-OCTANE ON PORASIl-C
PROGRAM: 50°C-3 MINUTES 6°/MINUTE TO I70"C
DETECTOR: ELECTROLYTIC CONDUCTIVITY
12	14	16
RETENTION TIME MINUTES
FIGURE 6 CHROMATOGRAM OF ORGANO HALIDES

-------
OPTIONAL
FOAM TRAP

- EXIT 1/
IN. O.D.
JI4MM. O.D.
INLET 1/4
IN. O.D.
1A IN. O.D. EXIT
10MM. GLASS FRIT
MEDIUM POROSITY
SAMPLE INLET
2-VMY SYRINGE VALVE
	 17CM. 20 GAUGE SYRINGE NEEDLE
6MM.O.D. RUBBER SEPTUM
10MM. O.D.
INLET
1/4 IN. O.D.
/
¦J
1/16 IN. O.D.
STAINLESS STEEL
/
13X MOLECULAR
SIEVE PURGE
GAS FILTER
*
PURGE GAS
FLOW CONTROL
• » • isrwi^r

-------
PACKING PROCEDURE
MULTIPURPOSE TRAP
OENERAL PURPOSETRAP
OLASS WOOL 5MM
GRADE IS
SILICA GEL
8CM
IENAX
15CM
3% OV-I »CM
GLASS WOOL
5MM
I
M
m
m
;f/
'd?
i®
4v
¦$;
'Wi
tw
Is
m
it
%.
m

GLASS WOOL 5MM
y3
TRAP INLET
TEN AX
23CM
3JS OV-1
GLASS WOOL
1CM
S MM
Ufl
M
tU
$
M
$
§t
'l.fy
4
M
A',
u
:S|
/ i.v*
$V;
'fe-i
:iix.
M
TRAP IN1ET
FIGURE 2 TRAP
CONSTRUCTION

-------
CARRIER OAS FLOW CONTROL
PRESSURE REGULATOR
UQUID INJECTION PORTS
PURGE GAS
FLOW CONTROL \
13X MOLECULAR
SIEVE FILTER
COLUMN OVEN
—CONFIRMATORY COLUMN
TO DETECTOR
ANALYTICAL COLUMN
HEATER CONTROL
PURGING DEVICE
NOTE: ALL LINES BETWEEN
TRAP AND GC
SHOULD BE HEATED
TO 80°C
FIGURE 3 PURGE-TRAP SYSTEM (PURGE-SORB MODE)

-------
CARRIER GAS FLOW CONTROL
PRESSURE REGULATOR
LIQUID INJECTION PORTS
PURGE GAS v
FLOW CONTROL X
13X MOLECULAR
SIEVE FILTER

COLUMN OVEN
	CONFIRMATORY COLUMN
TO DETECTOR
ANALYTICAL COLUMN
HEATER CONTROL
NOTE: ALL LINES BETWEEN
TRAP AND GC
SHOULD BE HEATED
TO 80°C
PURGING DEVICE
FIGURE 4 PURGE-TRAP SYSTEM (DESORB MODE)

-------
COLUMN: 0.2% CARBOWAX 1S00 ON CARBOPACK-C
PROGRAM: 60°C-3 MINUTES 8°/MINUTE TO 160°C
DETECTOR: ELECTROLYTIC CONDUCTIVITY S
ec
O
8	10	12	14
RETENTION TIME MINUTES
riounr c ri inrvn	A HA AC AD/^AKIAUAI IHPC

-------