Purge and Trap Analysis Using Fused Silica Capillary Column Gas Chromatography/ Mass Spectrometry BY F. A. Dreisch and T. 0. Munson* Central Regional Laboratory, Region 3 U.S. Environmental Protection Agency 839 Bestgate Road Annapolis, Maryland 21401 U. S. Environmental Protection Agency Environmental Science Center 701 Mapes Road Ft. Meade, MD 20755-5350 ------- *Current Address: Chemistry Research Unit Forensic Science Research and Training Center F.B.I. Academy Quantico, Virginia 22135 ------- BRIEF ABSTRACT Purge-and-trap GC/MS analysis using a fused-si1ica capillary column was compared to the EPA Method 624 which uses packed columns. The capillary column method was found comparable for spiked water samples and far superior for complex environmental samples. ------- ABSTRACT Purge-and-trap analysis using fused-silica capillary column gas chromatography/mass spectrometry (GC/MS) was evaluated for analysis of priority pollutant organics compared to the EPA Method 624 which uses packed columns. Using a packed column, our purge-and-trap GC/MS system generated precision data very similar to that reported by others (5.8% compared to 5.1%). Preliminary tests of a 30 m SE54 capillary column system, using the EPA Volatile Organic Performance Evaluation Samples WP006 and WP007 unknown test materials, indicates satisfactory precision (8.5%) and excellent accuracy (98.5 and 102 percent recoveries). Not only was the capillary column much faster than the packed column method (13 min. compared to 33 min.), it also overcame difficulties normally encountered with samples of excessive complexity, or very large concentration spread among components, or containing high concentrations of high boiling components. ------- INTRODUCTION Purge-and-trap gas chromatography/mass spectrometry (GC/MS) is a widely used technique for the measurement of volatile organic compounds in aqueous samples—particularly for the screening of environmental samples for the U.S. Environmental Protection Agency's (EPA's) list of volatile priority pollutants (1,2). We have analyzed several hundred such samples employing the technique essentially as described in the EPA recommended Method 624 (2), and while for most samples the method was quite satisfactory some samples presented serious problems. These problems can be classified as three types: 1.) excessive complexity--the eluting compounds overlap extensively making identification difficult; 2.) excessive concentration spread—with the lowest-level components just above the detection limit, some of the highest-level components cause very broad peaks obscuring many of the lowest-level components; and 3.) high concentrations of high boiling components—the gas chromatographic runs must be lengthened to allow the late peaks to elute, greatly extending the required analysis time. Others have reported measuring the EPA volatile priority pollutants by purge-and-trap GC/MS using high resolution capillary column chromatography with cryogenic trapping of the desorbed compounds onto the head of the capillary column (3,4). We have established this method in our laboratory.* We will describe our version of this technique in this report, the precision and accuracy of the technique when used for determining some of the EPA priority pollutants, and how the technique does overcome the difficulties we encountered when using Method 624 on certain samples. * A portion of this work was presented at the 16th Middle Atlantic Regional Meeting of the American Chemical Society, Newark, Delaware, April 21-23, 1982 ------- EXPERIMENTAL Materi als The GC/MS reference compound 4-bromofluorobenzene was obtained as the pure compound from PCR Research Chemicals, Inc., Gainesville, Fla. All of the other compounds were obtained as prepared standard solutions from Supelco, Inc., Bellefonte, Pa.: Purgeable A (0.2 g/L in Methanol) Purgeable B (0.2 q/L in Methanol) methylene chloride trichlorof1uoromethane 1,1-dichloroethene (E)-l,2-dichloroethene 1,1-dichloroethane 1,2-dichloroethane chloroform 1,1,1-trichloroethane carbon tetrachloride bromodichloromethane 1,2-dichloropropane (E)-l,3-dichloropropene trichloroethene (Z)-l,3-dichloropropene 1,1,2-trichloroethane benzene dibromochloromethane bromoform tetrachloroethene 1,1,2,2-tetrachloroethane chlorobenzene toluene ethylbenzene Purgeable C (0.2 q/L in Methanol) Internal Standards (20 g/L in Methanol) chloromethane bromochloromethane dichlorodifluoromethane 1,4-dichlorobutane bromomethane 2-bromo-l-chloropropane vinyl chloride chloroethane 2-chloroethyl vinyl ether (0.2 g/L in Methanol) Acrolein/acylonitrile (1 g/L in Water) ------- Other materials were obtained as follows: Tenax GC, 60/80 mesh (Applied Science Laboratories, Inc, State College, Pa.); activated coconut charcoal, 40/60 mesh (Analabs, Inc, North Haven, Ct.); and 1% SP1000 on 60/80 mesh Carbopack B (Supelco, Inc). Organic-free water was generated in the laboratory by passing water from a Super-Q System (Mi 11ipore Corp., Bedford, Massachusetts) through a cannister (all glass and metal) containing about 750 gms of activated charcoal (NUCHAR WV-L, 8/30 mesh, Westvaco, Covington, Va.). APPARATUS A Finnigan 3300 GC/MS (Finnigan Instruments, Sunnyvale, Ca.) coupled to an RDS Dual Computer Data System (RDS Nermag, Santa Clara, Ca.) was used for this work. The mass spectrometer was operated in the electron impact ionization mode (70 eV) using instrument settings similar to those suggested by Budde and Eichelberger (5). The GC/MS unit was modified by replacing the Finnigan 9500 GC with a Hewlett-Packard 5840A GC equipped with a capillary injector system (Hewlett-Packard Co., Avondale, Pa.). The packed column data were generated using a 2.4 m x 2.1 mm I.D. stainless steel column packed with 1% SP1000 on 60/80 mesh Carbopack B, and the capillary column data were generated using a 30 m x 0.25 mm I.D. SE-54 fused-silica capillary column (J&W Scientific, Rancho Cordoba, Ca.). When a packed column was used, the outlet of the column was connected ,to the mass spectrometer by a glass ------- jet separator enrichment device and a glass-lined stainless steel transfer line. The separator oven and the transfer line were kept at about 225°C. In order to connect the capillary column to the mass spectrometer, the glass jet separator was removed and the end of the capillary column passed from the GC oven into the separator oven, down the transfer line into the vacuum manifold with the end of the column positioned about 1 mm back from the plane of the opening into the ion volume. The purge-and-trap procedures and the necessary equipment have been described in great detail elsewhere (2,6), and only specific operating parameters and modifications will be included in this report. An automatic Hewlett-Packard 7675A Purge-and-Trap Sampler was mounted on t-he HP5840A GC with the heated, rigid portion of the sample line assembly extending through the top of the GC oven. When a packed GC column was used, the effluent from the sampler was routed directly to the GC column by connecting the end of the flexible, 1.6 mm (1/16 in) O.D. stainless steel tubing portion of the sample line assembly to the top of the GC column. The sampler trap needed to be desorbed with 15 to 30 ml/min gas flow. Since the capillary column needed a flow of 0.5 to 1.0 ml/min, the effluent from the sampler was connected to the capillary column via the capillary injector, with the injector adjusted to yield about a 20:1 split, when the capillary column was used. The flexible portion of the sample line assembly was routed back through the top of the GC oven and connected to a 23 cm length of 3.2 mm (1/8 in) O.D. teflon tubing fixed to a syringe needle which pierced the septum on the capillary injector. The trap on the sampler (4.5 mm I.D.) was packed with 4.5 cm of Tenax at the inlet end followed by 4.5 cm of activated charcoal, which has been found suitable for trapping all of the EPA priority pollutant volatiles (7). ------- Cryogenic trapping was necessary to achieve narrow peak widths on the early eluting compounds, particularly those compounds more volatile than methylene chloride. The portion of the fused-silica capillary just downstream of the injector was shaped into a loop by being taped to a rod extending down from the top of the GC oven. The configuration was such that, when an insulated 100 ml beaker full of liquid nitrogen was positioned around the loop, about 14 cm of the capillary column was immersed in the liquid nitrogen. For all of the purge-and-trap runs described in this report, a 5.0 ml sample volume was purged in the large (25 mm x 150 mm) sample tube using the "deep vortex" technique (8). An egg-shaped magnetic spin bar (20 mm long) in the bottom of the sample tube just below the terminus of the purge gas delivery tube was caused to rotate with sufficient speed that a deep vortex formed all the way to the spin bar. When a sample was purged, the purge gas passed over the surface of the sample but not through it, thus eliminating the problem of sample foaming. For the samples which were heated during the purging process, a small heated water bath was constructed out of a clear plastic food storage container (18.5 cm x 14.5 cm x 10.5 cm) with a tight-fitting lid (18.5 cm x 14.5 cm x 1.5 cm). A Braun Thermomix 1419 heating-stirring unit (SGA Scientific, Inc, Silver Spring, Md.) was inserted into one end of the water bath through a hole cut in the lid; another hole was cut through the lid at the other end to allow the bath to be raised up under the sample tube on the purge-and-trap device. , With the sample tube inserted completely to the bottom of the wat'er-bath, the magnetic mixer could be placed under the water bath to turn the stir bar in the sample tube. ------- The HP7675A Purge-and-Trap Sampler was designed with a requirement for fairly high-volume 70 psi air to rotate two pressure-activated valves and to cool the trap. Because our laboratory lacked centrally-supplied high-pressure air, the air requirement was initially satisfied using a pure air generator. After several inconvenient and costly failures of the air generating system, the air flow system on the 7675A Purge-and-Trap Sampler was restructured so that the pressure-activated valves were supplied 70 psi air from a compressed air cylinder, and the trap cooling system was supplied 15 psi air from the laboratory air system. Although, thus configured, the trap seemed to be cooled adequately between runs, additional cooling air was supplied to the exterior of the trap heating assembly using 6.4 mm (1/4 in) I.D. tubing to supply 15 psi laboratory air during the trap cooling cycle. Quality Control The performance of the GC/MS system was monitored through the daily use of a GC/MS reference compound (9). The instrument was tuned so that the purge-and-trap analysis of 50 ng of 4-bromof1uorobenzene (BFB) provided a spectrum for that compound which satisfied recommended ion abundance ratio criteria (10). This compound was used as one of the surrogate compounds which were spiked into each sample analyzed. In this manner, the tune of the instrument could be verified for each sample run. The BFB criteria were used as a "benchmark" reference for tuning the GC/MS system and monitoring its performance, rather than as a performance requirement which had to be slavishly satisfied prior to sample analysis. ------- Every purge-and-trap run (method blank, field blank, standards run, or sample) was spiked with a four-component standard spike: bromochloromethane, 1,4-dichlorobutane, 2-bromo-l-chloropropane, and 4-bromof1uorobenzene, to give concentrations of 20, 20, 20, and 10 ug/L respectively. The ion current for a particular ion for each compound (128, 77, 55, and 176, respectively) was measured for each purge-and-trap run. On the one hand, these data were used as a diagnostic tool to monitor the performance of the entire purge-and-trap GC/MS system in terms of instrument response, assuming that the data from these surrogate compounds would be applicable to all of the compounds of interest. On the other hand, one of the compounds (2-bromo-l-chloropropane) was used as the internal standard for quantitation of the other three compounds (and all other compounds detected) when compared to their values in the standards run. The calculated mean concentrations recovered for a particular sample set, and the relative standard deviations from the mean, provided precision and accuracy information for the entire analysis (from filling the syringe through processing the data) for those three surrogate compounds, and by inference, for all of the compounds of interest. The mean concentrations and the relative standard deviations were reported in the Quality Control Summary provided with each analytical report. Reagent blanks (organic-free water) were run both at the beginning of each day to demonstrate that the purge-and-trap GC/MS system was sufficiently free of internal contamination, and whenever it was necessary to measure possible carry-over of compounds from one run to the next (e.g., after a sample containing very high concentrations, or after a standards run). A field blank consisting of organic-free water was transported to the sampling site, taken back to the laboratory with the samples, and stored with the samples until the time of analysis. It was then analyzed with each set of samples. ------- One of the samples from each sample set was analyzed in duplicate and the values reported as part of the Quality Control Summary. Occasionally sample duplicates (5 to 10% of the samples) were spiked with all of the priority pollutant purgeable standards, and the recoveries reported in the Quality Control Summary. QUANTITATION The only compounds which were routinely quantitated against standards run on our instrument were the EPA priority pollutant volatile organic compounds. Each compound was quantitated using the ion current for a specific ion (usually the ion recommended in Method 624) compared to a<- single standards run (one concentration) using the internal standards method as described in Method 624, with 2-bromo-l-chloropropane as the internal standard. All of the data in this report were generated using the RDS Automatic Search and Quantitation software program. In order to identify non-priority pollutant compounds, background corrected spectra were processed with the Library Search Routine contained in the RDS SADR Data Reduction Program. A library disk supplied by RDS Nermag (NBS Library Disk, containing the 35,000 spectra EPA/NIH data base in abbreviated form) was used as the on-line data base. Using the list of the ten best-fit compounds found in this manner as a guide, the background corrected spectrum was compared manually to the spectra displayed in the EPA/NIH Mass Spectra Data Base volumes (11). A compound was considered presumptively identified if all of the ions (including those with ion abundances less than 5%) in the published spectrum were present in the sample spectrum, and any additional ions in the sample spectrum could be accounted for. ------- The concentrations of identified compounds for which standards had not been run were grossly estimated using the average of the responses of the compounds in the four component spike. PROCEDURE Aqueous samples were collected as specified in Method 624 (2) and stored refrigerated together with the appropriate field blanks until the time of analysis. Four replicates of each sample were collected routinely in case replicate analysis was necessary at some later date. At the time of analysis, the sample was poured into the barrel of a 5 ml gas-tight syringe equipped with a two-way valve (in the open position) and a 15 cm 20 gauge needle. With the barrel of the syringe filled to overflowing, the plunger was inserted, depressed to the 5.0 ml mark and the valve closed. If a laboratory duplicate of the sample were run, at this point a second syringe assembly would have been loaded with the same sample. The appropriate standard solution was added to each sample by removing the needle from the sample syringe assembly and passing the needle of the spiking syringe through the two-way valve into the sample contained in the sample syringe assembly. Every sample was spiked with 5.0 ul of the four component standard spiking solution (an organic-free water solution containing 20 mg/L of three components and 10 mg/L of 4-bromofluorobenzene). The needle was replaced on the syringe assembly and the sample transferred to the HP7675A sample tube containing the magnetic stir bar, care being taken not to atomize the sample. The loaded sample tube was then mounted on the HP7675A, the deep vortex established, and the purge-and-trap cycle started. ------- Fish samples were frozen with solid CO^ in the field as they were collected and stored frozen until they were prepared for analysis. Each fish was cut into cubes with a band saw and the cubes coarsely ground with a sausage grinder. The resulting icy slush was stirred and replicate subsamples (about 5 gm each) stored frozen in 40 ml VOA vials (Pierce No.13075, Pierce Chemical Co., Rockford, 111.). At the time of analysis, ice-cold organic-free water (20 ml) was added to the vial of frozen ground fish and the vial shaken vigorously until the granules of fish were suspended in the water. The slurry was then sonicated as described by Easley, et al. (12), with 20 1 of the 4 component standard spike added prior to the sonication. A 5 ml portion of the sonicated slurry was measured with an ice-cold 10 ml graduated cylinder, poured into a HP7576A sample tube and analyzed according to the procedure for aqueous samples. For all of the data presented in this report, the purge-and-trap/desorb sequence was the same: a 1 min pre-purge period during which cooling air was passed over the trap; an 11 min purge-and-trap period during which a 40 ml/min flow of helium was passed sequentially through the purge vessel and trap; a 4 min desorb period during which the trap was heated to 180°C and backflushed with a 15 ml/min stream of helium (20 ml/min for packed column runs); and a 9 min vent period during which the trap was baked at 230°C while being backflushed with 40 ml/min helium. When the packed column (1% SP1000) was used, the column temperature was 45°C during the 4 min desorb period while the compounds were being stripped from the trap and carried to the GC column; then the temperature was increased to 225°C at 10°C/min,' and maintained at 225°C for 10 min. The temperature was further increased at 20°c/min to 245°C and maintained for 4 min to sweep late-eluting compounds from the column, resulting in a total GC run time of 37 min. ------- When a capillary column was used, the gas stream carrying the desorbed compounds was split approximately 20:1 in the capillary injection port. The cryogenic trap was placed around the column about 1 min prior to the start of the desorb period and removed exactly at the end of the desorb period. Two GC temperature programs were used with the capillary column—a short program when late-eluting peaks were not expected to occur, and a longer program when they were. For the short program, the column was held at 30°C for 5.5 min (measured from the start of trap desorb), heated at 2°C/min for 4.5 min, and heated at 10°C/min to a final temperature of 110°C. The total GC run time was 17.5 min. For the longer program, the temperatures were programmed exactly the same for the first 15 min, then the temperature was increased to 150°C at 20°C/min and held for 5 min giving a total run time of 22.5 min. For the packed column purge-and-trap runs, the data collection by the GC/MS data system was initiated coincidentally with the start of the desorb phase. A split mass range (20-27; 33-260 amu) was scanned at 9 msec/amu with a resultant scan rate of 2.4 sec/scan. For the capillary runs, data collection was initiated 3.0 min after the start of the desorb phase. The same split mass range was scanned at 0.86 sec/scan for the initial capillary runs---the fastest scan rate attainable by the GC/MS data system interface at that time. Subsequently, the interface was modified by RDS to decrease the settling time, and a scan rate of 0.63 sec/scan was used. ------- RESULTS AND DISCUSSION Table 1 shows the results of a precision study for our purge-and-trap GC/MS system when operated using a packed column. The mean relative standard deviation (RSD) of 5.8 percent compares favorably with the values reported by Olynyk, et al. (10) of 5.2, 5.4, and 5.0 percent for selected groups of these compounds. Their tests did not include six of the most volatile priority pollutant compounds (chloromethane, bromomethane, vinyl chloride, dichlorodifluoromethane, chloroethane, and trichlorofluoromethane), and two compounds of quite poor purging efficiencies (acrolein and acrylonitrile). Calculating a mean RSD from the data in Table 1 excluding these compounds yields 5.1 percent, which is even closer to the reported values. Figure 1 shows the extracted ion current profile (EICP) from a fused-silica capillary column purge-and-trap GC/MS run of all of the priority pollutant purgeable organics compared to the EICP obtained when a packed column was used. The tremendous resolving power of the capillary column is obvious--the capillary peaks are so narrow that during two-thirds of the run time, no peaks were eluting. The potential exists for adequately resolving mixtures much more complex than this 35 component mixture. The increased resolution is achieved using a much shorter run time for the capillary run, only about 13 mins to elute the last compound compared to about 33 mins for the packed column (the time axis on the EICP for the packed column includes the 12 min purge time). Better separation of the individual components in the mixture can be achieved using a longer capillary run time (and flatter temperature ramps), but because one can distinguish between overlapping compounds using the specific ions recorded by the mass spectrometer, we chose to trade increased separation for shorter run times. ------- Table 2 presents data which allow an estimation of the precision of the capillary purge-and-trap method for some of the priority pollutant purgeable organic compounds. The mean RSD of 8.5 percent which can be calculated from these data represents only a rough estimate of the precision of this method for the priority pollutant purgeables because fewer than one third of the compounds in the mix were represented in the test, and only three runs were performed. Table 3 presents precision data from the duplicate analysis by capillary purge-and-trap GC/MS of a "real world" sample from a steel mill treatment lagoon. In addition to the compounds shown in Table 3, this sample contained 15 non-priority pollutant volatile organic compounds in the 1-100 ug/L range. Table 4 presents data generated during tests of the accuracy of the capillary purge-and-trap GC/MS method using the EPA Volatile Organic Performance Evaluation Samples WP006 and WP007 as unknown test materials- As might be expected from the good mean recoveries achieved (98.5 and 102 percent), all of the values fell within the reported 95% confidence intervals (13). Table 5 shows the use of the surrogate standards to provide both precision and accuracy information for a data set consisting of 16 capillary purge-and-trap GC/MS runs accumulated during the analysis of 6 steel mill samples. The mean precision and accuracy values were quite satisfactory considering the nature of the samples. A closer examination of the surrogate data shows reduced recoveries in 2 of the 6 steel mill samples. Table 6A shows the concentrations of the spiked surrogate standards which were calculated using the internal standard method, and Table 68 shows the raw data which were used to generate the values shown in Table 5A. The internal standard method yielded very acceptable quantitation of the surrogate compounds in all 6 ------- of the samples even though the raw data indicate very poor recovery of the surrogate compounds from sample 6, and somewhat diminished recovery from sample 4. These poor recoveries represent "matrix effects" as discussed by Olynyk, et al. (10). Because both of these samples contained more than 50 compounds, many at quite high concentrations (samples 6 and 4 are estimated by this purge-and-trap analysis to contain 0.2-2% and 0.01-0.03% total organics by weight, respectively), the surrogate compounds were probably more soluble in these sample matrices than in the less contaminated samples or the organic-free water of the blanks and spikes, and therefore, purged less efficiently. One might guess from these results that the internal standard method would adequately correct for this type of solubility matrix effect for a broad range of organic compounds which are sparingly soluble in water. Purge-and-trap analysis of compounds less volatile than 4-bromofluorobenzene is not generally considered practical using the packed columns suggested in Method 624 (2) due to the excessively long elution times involved. That such compounds do purge-and-trap is clearly evident from the late-eluting peaks which frequently occur as lumps in subsequent runs. The use of capillary purge-and-trap greatly extends the range of compounds that can effectively be analyzed by the purge-and-trap technique. Table 7 presents a list of the compounds which we detected using fused-silica capillary purge-and-trap GC/MS. ------- Figure 2 presents an EICP (from the purge-and-trap analysis of a fish tissue sample) which illustrates the power of the method described in this report. This sample run displays: 1.) the excessive complexity which could not have been adequately handled by a packed column; 2.) the broad range of concentrations which would have lead to a loss of valuable information on a packed column (the small peak just after scan 800 represents about 100 ng); and the high concentration of late eluting peaks (everything after scan 1050) which would have necessitated long delays between packed column runs. ------- Table 1 . Packed Column Method Precision Compound3 % RSDb Compound9 % RSDb chloromethane 5.9 1,2-dichloropropane 4.2 bromoethane 8.8 (E)-1,3-dichloropropene 2.7 vinyl chloride 8.2 trichloroethene 5.5 di chlorodif1uoromethane 8.2 benzene 6.5 chloroethane 7.5 (Z)-l,3-dichloropropene 3.3 methylene chloride 14 1,1,2-trichloroethane 4.3 acrolei n 8.2 dibromochloromethane 4.8 acryloni trile 5.4 2-chloroethyl vinyl ether 5.8 trich1 orof 1uoromethane 11 2-bromo-l-chloropropane (I.S.) c 1,1-dichloroethene 7.6 bromoform 5-4 bromochloromethane 6.4 1,1,2,2-tetrachloroethane 1.3 1,1-dichloroethane 6.1 tetrachloroethene 4.2 (E)-l,2-dichloroethene 8.4 1,4-dichlorobutane 1.2 chloroform 6.2 toluene 4.7 1,2-dichloroethane 5.5 chlorobenzene 4.9 1,1,1-tri chloroethane 3.0 ethylbenzene 4.7 carbon tetrachloride 4.7 4-bromof1uorobenzene 3.1 bromod i ch1oromethane 4.8 mean RSD = 5.8% a. All of the compounds were at 20 ug/L except for the following: chloromethane, bromomethane, vinyl chloride, dichlorodifluoromethane, chloroethane at 40 ug/L; acrolein and acrylonitrile at 400 ug/L; and 2-chloroethyl vinyl ether at 100 ug/L. b. Percent relative standard deviation calculated using the data from 5 standard runs c. This compound was used as the internal standard for the calculations. ------- Table 2. Precision Study: WP007 Compound chloroform 1,2-dichloroethane ],1,1-trichloroethane carbon tetrachloride tri chloroethene bromodichloromethane dibromochloromethane tetrachloroethene chlorobenzene bromoform methylene chloride mean RSD = 8.5 + 5.5 Concentration Measured (ug/L) Sample # . Rep. 1 Rep. 2 Rep. 3 % RSD 1 7.8 6.4 7.0 9.9 2 54 49 52 4.9 1 3.5 2.8 3.6 13 2 68 64 66 3.0 1 35 30 30 9.1 2 5.4 4.9 4.9 5.7 1 39 32 31 13 2 9.6 8.8 8.8 5.1 1 7.7 6.0 6.0 15 2 25 22 14 28 1 2.1 1.9 1.8 7.9 2 16 16 16 0 1 5.1 4.5 4.9 6.3 2 22 19 19 8.7 1 17 15 14 10 2 36 32 31 8.0 1 5.3 4.8 5.6 7.7 2 26 23 25 6.2 1 32 27 30 8.5 2 2.7 2.6 2.3 8.2 1 16 14 15 6.7 2 3.6 3.8 3.8 3.1 ------- Table 3. Precision in the Analysis of a Sample From a Steel Mill Treatment Lagoon Concentration in ug/L Compound Sample Duplicate % RSD methylene chloride 18 18 0 1,1-dichloroethene 0.75 1.1 27 chloroform 7.4 7.5 0.95 1,1,1-trichloroethane 34 34 0 1,2-dichloroethane 7.4 7.5 0.95 benzene 1.4 1.8 18 trichloroethene 7.0 6.4 6.3 bromodichloromethane 1.2 1.3 5.7 toluene 28 28 0 tetrachloroethene 3.7 4.0 5.5 chlorobenzene 1.8 1.9 3.8 ethyl benzene 5.1 4.7 5.8 mean RSD = 6.2 + 8.3 ------- Table 4. Accuracy Study A. WP006 Single Analysis Compound Sample No. Result3 True Value9 % Recovery 1,2-dichloroethane 1 44.8 47.8 93.7 2 13.9 12.8 109 chloroform 1 67.7 74.6 90.8 2 11.0 11.2 98.2 1,1,1-trichloroethane 1 21.4 24.1 88.8 2 3.30 3.21 103 trichloroethene 1 28.0 28.3 98.9 2 3.20 3.33 96.1 carbon tetrachloride 1 46.2 60.2 76.7 2 2.90 3.01 96.3 bromodichloromethane 1 12.5 12.0 104 2 36.1 34.2 106 dibromochloromethane 1 48.1 42.8 112 2 8.10 8.56 94.6 bromoform 1 20.5 18.2 113 2 4.90 4.05 121 methylene chloride 1 6.7 8.37 80.0 2 32.2 33.4 96.4 chlorobenzene 1 37.0 41.4 89.4 2 13.0 13.8 94.2 tetrachloroethene 1 9.60 9.12 105 2 61.0 60.8 100 mean recovery = 98.5 + 10.6 aAll concentration values in ug/L ------- Table 4. Accuracy Study (con't) B. WP007 Multiple Analysis (3 runs) Compound Sample No. Result9 True Value3 % Recovery 1,2-dichloroethane 1 3.30 3.19 103 2 66.3 63.8 104 chloroform 1 7.07 7.46 94.8 2 51.6 55.9 92.3 1,1,1-trichloroethane 1 32.1 32.1 100 2 5.07 4.82 105 trichloroethene 1 6.56 6.66 98.5 2 23.2 25.0 92.8 carbon tetrachloride 1 34.1 36.1 94.5 2 9.05 9.04 100 bromodichloromethane 1 1 .94 1.71 113" 2 15.8 15.4 103 dibromochloromethane 1 4.84 4.28 113 2 20.0 19.3 104 bromoform 1 29.9 24.3 123 2 2.54 2.03 125 methylene chloride 1 14.9 16.7 89.2 2 3.74 3.34 112 chlorobenzene 1 5.24 5.52 94.9 2 24.3 27.6 88.0 tetrachloroethene 1 15.4 15.2 101 2 32.8 33.4 98.2 mean recovery = 102 + 9.69 aAll concentration values in ug/L ------- Table 5. Surrogate Standard Precision and Accuracy* Concentration Mean Recovered Compound Spiked (ug/L) Concentration (ug/L) % Recovered % RSD bromochloromethane 20.0 20.9 104 5.42 1,4-dichlorobutane 20.0 19.9 99.5 4.50 4-bromofluorobenzene 10.0 10.1 101 7.55 *These data are drawn from 16 runs spread over a two-day period; 8 of the runs were samples (6 sample runs plus 2 duplicate sample runs) from a steel mill, the remainder were the associated blanks and standards runs. ------- Table 6. Recovery of Surrogate Standards3 A. Expressed as Concentration ( Sample No. BCM 1 21.6 2 20.7 3 20.5 4 21.1 5 21.9 6 22.9 9/L) BCP DCB BFB (20) 20.2 9.64 (20) 20.4 10.0 (20) 19.0 9.94 (20) 20.7 9.34 (20) 19.0 11.1 (20) 20.4 11.7 mean 21.4 (20) 20.0 10.3 % RSD 4.13 --- 3.77 8.88 % Recovery 107 — 100 103 B. Expressed as Area Counts Sample No. BCM 1 340.0 2 348.2 3 335.6 4 303.1 5 348.2 6 147.2 BCP DCP BFB 696.3 735.2 278.5 704.5 823.3 299.0 686.1 747.5 290.8 606.2 696.3 274.4 667.6 729.1 315.4 270.3 315.4 135.2 meanb 335.0 672.1 746.3 291.6 % RSDb 5.57 5.85 6.30 5.66 a. BCM = bromochloromethane, 128 amu; BCP = 2-bromo-l-chloropropane, 77 amu; DCB = 1,4-dichlorobutane, 55 amu; BFB = 4-bromofluorobenzene, 176 amu. b. The values for sample 6 were not included in the calculations of the mean values or the % RSD values. ------- Table 7. Compounds Identified Using Purge-and-Trap Capillary Column GC/MSa an Number Compound Identified 274 sulfur dioxide 277 hydrogen sulfide 278 chloromethane 281 vinyl chloride 286 butane 287 bromomethane 290 methanethiol 290 chloroethane 298 ethanol 303 trichlorofluoromethane 303 acetone 305 diethyl ether 313 ethanethiol 317 1,1-dichloroethene 317 acrylonitrile 318 methylene chloride 321 thiobismethane 322 1,1,2-trichloro-l,2,2-trifluoroethane 328 carbon disulfide 338 (E)-1,2-dichloroethene 347 2-metboxy-2-methylpropane 353 thirane 354 1,1-dichloroethane 367 2-methyl-2-propanethiol 367 2-butanone ------- Table 7. Compounds Identified Using Purge-and-Trap Capillary Column GC/MSa(con1t) an Number Compound Identified 367 diisopropyl ether 368 pentane 373 2-butanol 379 (Z)-l ,2-dichloroethene 380 ethyl acetate 383 1-propanethiol 383 bromochloromethane 383 chloroform 388 (methylthio)ethane 401 tetrahydrofuran 415 methylthirane 420 2-methyl-1-propanol 420 1,1,1-tri chloroethane 427 1,2-dichloroethane 432 hydrochloric acid 433 1-bromo-1-chloroethane 438 cyclohexane 443 carbon tetrachloride 444 benzene 458 2-(methylthio)propane 458 2-butanethiol 494 2-pentanone 508 3-pentanone 509 1,2-dichloropropane 511 heptane ------- Table 7. Compounds Identified Using Purge-and-Trap Capillary Column GC/MSa(con't) Scan Number Compound Identified 513 dibromomethane 515 trichloroethene 524 bromodichloromethane 525 1 , 1 '-thiobisethane 531 1,4-dioxane 552 methylcyclohexane 585 2-chloroethyl vinyl ether 601 (Z)-1,3-dichloropropene 605 4-methyl-2-pentanone 620 dimethyl disulfide 661 toluene 661 (E)-l,3-dichloropropene 675 1,1,2-trichloroethane 688 2-bromo-l-chloropropane 691 5-hexen-2-one 712 dimethylcyclohexane (isomer) 713 3-hexanone 730 dibromochloromethane 745 dimethylcyclohexane (isomer) 765 dimethylcyclohexane (isomer) 769 tetrahydrothiophene (T)*3 773 tetra'chloroethene 790 butyl acetate 835 ethylcyclohexane (isomer) 844 ethyldimethylcyclohexane (isomer) ------- Table 7. Compounds Identified Using Purge-and-Trap Capillary Column GC/MSa(con1t) Scan Number Compound Identified 864 chlorobenzene 877 2-methyl tetrahydrothiophene (7)13 895 ethylbenzene 915 xylene (isomer) 950 bromoform 958 2,5-dimethyl tetrahydrothiophene (isomer) 968 2,5-dimethyl tetrahydrothiophene (isomer)- 971 xylene (isomer) 975 1 ,3,5,7-cyclooctatetraene (7)13 990 nonane 1007 1,4-dichlorobutane 1020 1,1,2,2-tetrachloroethane 1033 1,2,3-trichloropropane 1041 4-bromofluorobenzene 1061 bromobenzene 1106 chloromethylbenzene (isomer) 1112 propyl benzene 1142 trimethylbenzene (isomer) 1148 dimethyltrisulfide 1171 benzonitrile 1182 6-methyl-5-hepten-2-one 1189 trimethylbenzene (isomer) 1199 decane 1211 butylbenzene ------- Table 7. Compounds Identified Using Purge-and-Trap Capillary Column GC/MSa(con't) Scan Number Compound Identified 1215 1,3-dichlorobenzene 1232 1 ,4-dichlorobenzene 1237 trimethylbenzene (isomer) 1258 2,3-dihydroindene 1271 lH-indene 1273 1-propynylbenzene 1282 1,2-dichlorobenzene 1289 diethylbenzene (isomer) 1294 tetramethylbenzene (isomer) 1304 methylphenol (isomer) 1307 methylpropylbenzene (isomer) 1324 dimethylethylbenzene (isomer) 1334 dimethylethylbenzene (isomer) 1344 methylphenol (isomer) 1345 nitrobenzene 1359 methylbenzoate 1379 tetramethylbenzene (isomer) 1386 tetramethylbenzene (isomer) 1408 3,5-dimethyl-1,2,4-trithiolane (isomer) 1418 3,5-d'imethyl-1,2,4-trithiolane (isomer) 1446 1,2,3,4-tetrahydronaphthalene 1478 naphthalene 1479 trichlorobenzene (isomer) ------- Table 7. Compounds Identified Using Purge-and-Trap Capillary Column SC/HSa{cori't) Scan Humber Compound Identified 1523 trichlorobenzene (isomer} 1S95 1,2,314-tetrahydromett)ylnaphthaleri6 (isomer) 1644 methyl naphthalene ("isomer) l£?3 methylncphthalene {isomer) 1718 tetrachlorobenzene (isomer) 1 SO? tetrachlorobenzene (isomer) a. Tae non-priority pollutant compounds otter than the cooipouids listed fn Table 1) were identified with a computer-based matching system using the EFA/lflH 'library of spectra- fa. The parenthetical symbol indicates that the reference spectrum had some minor fragments not found in the sample spectruw. ------- REFERENCES 1. Sampling and Analysis Procedures for Screening of Industrial Effluents for Priority Pollutants. U.S. Environmental Protection Agency, Environmental Monitoring and Support Laboratory, Cincinnati, Ohio, March, 1977, revised April, 1977, pp. 1-15. 2. Guidelines Establishing Test Procedures for Analysis of Pollutants, Proposed Regulations; Purgeables-Method 624. Fed. Reg. 44: 69532-69539 (1979). 3. F. DeWalle, and E. Chian. Presence of Priority Pollutants in Sewage and Their Removal in Sewage Treatment Plants: Grant 806102, S. Hannah, Project Officer. U.S. Environmental Protection Agency, Municipal Environmental Research Laboratory, Cincinnati, Ohio (Draft). 4. A. R. Trussel, J. G. Moncur, F-Y Lieu and L. Y. C. Leong. Simultaneous Analysis of All Five Organic Priority Pollutant Fractions. J.. High Res. Chrom. Chrom. Commun. 4: 156-163 (1981 ). 5. W. L. Budde and J. W. Eichelberger. Organics Analysis Using Gas Chromatography/Mass Spectrometry. Ann Arbor Science Publishers, Inc., Ann Arbor, Michigan, 1979, p. 12. 6. Operating and Service Manual: Purge and Trap Sampler 7675A. Hewlett Packard Co., Avondale, Pa., (1979). ------- 7. D. F. Bishop. GC/MS Methodology for Priority Organics in Municipal Waste Water Treatment, EPA 600/2-80-196, U. S. Environmental Protection Agency, Municipal Environmental Research Laboratory, Cincinnati, Ohio (1980). 8. J. Poole and W. D. Snyder. A Technique for the Reduction of Foaming in Purge and Trap Samples, Application Note AN 228-13, Hewlett Packard Co., Avondale, Pa. 9. J. W. Eichelberger, L. E. Harris and W. L. Budde. Reference Compound to Calibrate Ion Abundance Measurements in Gas Chromatography-Mass Spectrometry Systems, Anal. Chem. 47: 995-1000 (1975). 10. P. Olymyh, W. L. Budde and J. W. Eichelberger. Simultaneous Qualitative and Quantitative Analyses. I. Precision Study of Compounds Amenable to the Inert Gas-Purge-and-Trap Method. J. Chrom. Sci. 19: 377-382 (1981). 11. S. R. Heller, G. W. A. Milne, editors. EPA/NIH Mass Spectral Data Base, Volumes 1-4, 1978; Supplement 1, 1980. U.S. Superintendent of Documents, Government Printing Office, Washington, D.C. 12. D. M. Easley, R. D. Kleopfer, and A. M. Carasea. The Analysis of Volatile Organic Compounds in Fish. J. Assoc. Off. Anal. Chem. 64: 653-656 (1981). 13. P. Britton, Environmental Monitoring and Support Laboratory, U. S. Environmental Protection Agency, Cincinnati, Ohio, Personal Communication, 1981. ------- Acknowledgement The authors wish to thank Janet Roberson for diligently assembl this manuscript. ------- Tt^ocr: . JL TJIL£= PURCEMLE SIANDARDS ON 3011 SE-54 FS CAPILLARY COLUMN fJl£: OB .DA ACQUISITION IAfil£= CAPTD3 .AT DATE: 1/2171981 100Z= 3260416 XI 50.00 - 260.00 TIME- " 2^04 3=03 4=14 6:24 1:29 8:34 9=39 TITLE: PURGEAfiLE STANDARDS ON PACKED COLUMN SP-1000/CARB0PACK 0 FILE: 29 .DA ACQUISITION TABLE: PURGE .AT DATE: 4/3/1981 50.00 - 260.00 TIME ------- TJIi£ - nsfi SAflPlt 6Y PURSL AND TRAP FILL- 35 ACGUJSIIiOJt TABL£ HAfl£= CAPTM mi- S/30/138J r>/?£/t,c i' ^ *• . / V ^ 1007. 3G")llSSf x: SO.00 - 260.0) SCAf( 1000 ------- Purge and trap analysis using fused silica capillary column gas EJDM OD79 C45J60 ------- |