v-,EPA
          United States
          Environmental Protection
          Agency
             Office of Research and
             Development
             Washington DC 20460
EPA/600/R-99/041
June 1999
Laser Fluorescence EEM
Probe for Cone Penetrometer
Pollution Analysis

-------
                                                                     EPA/600/R-99/041
                                                                            June 1999
LASER FLUORESCENCE EEM PROBE FOR CONE PENETROMETER POLLUTION ANALYSIS
                                      by
                                Jonathan E. Kenny
                       Tufts University, Chemistry Department
                               Medford, MA, 02155
                         Cooperative Agreement CR-821856
                                 Project Officer
                                  Bob K. Lien
                   Subsurface Protection and Remediation Division
                      National Risk Management Research Lab
                                 Ada, OK, 74820
                   National Risk Management Research Laboratory
                        Office of Research and Development
                       U.S. Environmental Protection Agency
                               Cincinnati, OH 45268

-------
                                                                      EPA/600/R-99/041
                                                                              June 1999
LASER FLUORESCENCE EEM PROBE FOR CONE PENETROMETER POLLUTION ANALYSIS
                                         by
                                  Jonathan E. Kenny
                          Tufts University, Chemistry Department
                                 Medford, MA, 02155
                            Cooperative Agreement CR-821856
                                    Project Officer
                                     Bob K. Lien
                      Subsurface Protection and Remediation Division
                         National Risk Management Research Lab
                                   Ada, OK, 74820
                      National Risk Management Research Laboratory
                           Office of Research and Development
                          U.S. Environmental Protection Agency
                                 Cincinnati, OH 45268

-------
                                             NOTICE
The U. S. Environmental Protection Agency through its Office of Research and Development partially
funded and collaborated in the research described here under Cooperative Agreement No. CR-821856 to
Tufts  University.  It  has  been  subjected to the Agency's peer and  administrative review and has been
approved for  publication as an EPA document.  Mention of trade names or commercial products does not
constitute endorsement or recommendation for use.

All research  projects  making  conclusions  or  recommendations  based  on  environmentally  related
measurements and funded by the Environmental protection Agency are required to participate in the Agency
Quality Assurance Program.  This project was conducted under an approved Quality Assurance Project Plan.
The  procedures specified in this  plan were  used  without exception.   Information  on the plan  and
documentation of the quality assurance activities and results are available from the Principal Investigator.

-------
                                          FOREWORD
The U.S. Environmental Protection Agency is charged by Congress with protecting the Nation's land, air,
and water resources.  Under a mandate of national environmental laws, the Agency strives to formulate and
implement actions leading to a compatible balance between human activities and the ability of natural
systems to support and nurture life.  To meet these mandates, EPA's research program is providing data and
technical  support for solving environmental problems  today and  building a  science knowledge  base
necessary to manage our ecological resources wisely, understand how pollutants affect our health, and
prevent or reduce environmental risks in the future.

The National Risk Management Research Laboratory (NRMRL) is the Agency's center for investigation of
technological and management  approaches for reducing risk  from threats  to human health and the
environment. The focus of the Laboratory's research program is on methods for the prevention and control
of pollution to air, water, and subsurface resources; protection of water quality in public water systems;
remediation of contaminated  sites and ground water; and prevention and control of indoor air pollution.
The goal of this research effort is to catalyze development and implementation of innovative, cost-effective
environmental technologies; develop scientific and engineering information needed by EPA to  support
regulatory and policy decisions; and provide technical support and information transfer to ensure effective
implementation of environmental regulations and strategies.

This project represents a cooperative effort between the Tufts University and the Environmental Protection
Agency. This report summarizes research, development, and field test of a multi-channel fiber optic LIF
(laser induced fluorescence) EEM (excitation emission  matrix) instrument for CPT  (cone penetrometer
technology) field deployment. The findings of this project are directly applicable to the site characterization
and subsurface pollution analysis at hazardous waste sites.
                                                 Clinton W. Hall, Director
                                                 Subsurface Protection and Remediation Division
                                                 National Risk Management Research Laboratory
                                                in

-------
                                           PREFACE

For the past fifteen years, our research group has been interested in  developing  new technology  for
monitoring groundwater and soil quality, using lasers and fiber optics to perform in situ measurements.  Our
original application was  surface waters and  monitoring  wells, and, after enough laboratory studies to
convince ourselves of the feasibility of the laser-induced fluorescence  approach, we performed a very
limited number of field tests of a first-generation instrument.  This work was published in Analytical
Instrumentation in 1987, and we returned to the laboratory to develop new, more powerful instrumentation
that would develop some of the potential of laser-induced fluorescence that the first-generation instrument
had left unexplored.

We  had  no particular plans to  return  to  the  field,  but  were  content  to  develop field-adaptable
instrumentation in the relative comfort of the laboratory for the  next few years, especially when a
commercial instrumentation company showed an interest in our work. All this changed in early 1993 when
we were invited to an army-sponsored conference on cone penetrometer technology as a way to put fiber-
optics-based probes into the subsurface, an idea we had acknowledged early on as having economic merit,
because it would save the cost of digging wells.  We also received a request for proposals from EPA about
the same  time, so, still  unsure  about our interest in returning to the field, we decided to submit a
preproposal, attend the conference, and see what would happen.

I sent my postdoctoral fellow  of four years,  Todd Taylor,  to the conference, and he came back highly
enthused.  We were invited to submit a full proposal, and a few months later, we were back in the business
of field analytical chemistry, with the EPA Robert S. Kerr Environmental Research Laboratory  (RSKERL)
in Ada, Oklahoma as partners.

Unfortunately for the group, Todd Taylor found a permanent job, and  left the group just as work was
getting started. This  made it very challenging for the  two  new people, graduate  student Sean Hart and
postdoctoral fellow Jie Lin, to finish designing and constructing an instrument that  had been conceived by
Taylor.  Adding to the challenge was the fact that the EPA's new cone penetrometer (CPT)  vehicle would
not come to Tufts for an orderly installation, but, it turned out, had to be assembled in two stages, the first in
Ada, and the second a few weeks later, on site at Hill Air Force Base in Utah.

After this exciting experience,  the instrument  returned to  the lab and saw no field work the second year,
because of the peculiarities of the EPA team's travel  budget.  We finally found an opportunity, using
funding available from the Northeast Hazardous Substances  Research Center through the Tufts University
Center for Field Analytical Studies and Technology (CFAST), to bring the EPA CPT team  and vehicle to
the greater Boston area in June of the final year of the project.  The system was installed  and demonstrated
                                                IV

-------
at Hanscom Air Force Base, in Bedford, Massachusetts through the help of our Air Force coordinator at the
site, Tom Best.  Tom familiarized us with the site, shared his data and ideas, and arranged for us to receive
whatever other assistance we needed, including storage of the vehicle at the end of the study.   Over the next
three months, our field schedule provided us with all the excitement we had experienced that first summer,
and lacked that long second  year.  After field work  under EPA sponsorship at  Elizabeth City,  North
Carolina, we returned to Hanscom for a second visit, thanks in part to funding for our EPA team's travel
expenses provided by Tom Best and Hanscom.  Since the field work ended in  September, 1996, we have
been dividing our time between data analysis, final improvements to the  instrument, and preparation of
reports.  It has been a very productive return to  the field for our research  group, and we look forward to
continued collaboration with our friends at RSKERL in Ada.

-------
                                           ABSTRACT

A fiber optic LIF (Laser induced fluorescence) EEM (Excitation emission matrix) instrument for CPT
deployment has been successfully developed and field tested.  The system employs a Nd:YAG laser and
Raman shifter as a rugged field portable excitation source.  This excitation source simultaneously produces
> 20 beams of differing wavelength in the UV that can be selectively chosen for launch into fiber optics.
The system uses a pair of silica-clad-silica optical fibers for each of its channels (eight of which were used
in the work described), one to deliver excitation light from the  multi-wavelength laser excitation source to
the sample, and the other to conduct contaminant fluorescence to a grating spectrograph which utilizes a
charge-coupled device detector  to  record fluorescence intensity as a function of both excitation  and
emission wavelength. The system has automatic data acquisition  (depth and LIF) and some real time data
analysis capability.  The  instrument has undergone  several different calibrations  including wavelength,
analyte concentration, and standard compound fluorescence as a function of incident energy. Many sample
types have been characterized in the laboratory including single components on sands of different particle
sizes and clay, and fuel mixtures in solution and on sands.

This system has been installed in a CPT vehicle and has undergone four field tests at three different sites.
The  first  field test was  at Hill  AFB in Utah; hardware  failure  and low contaminant levels prevented
successful demonstration of the technology. The second field test was at Hanscom AFB in Massachusetts,
where we achieved our goals of instrument characterization and demonstration.  The third field test, at the
U.S.  Coast Guard Station in Elizabeth City,  North Carolina, was intended as a demonstration of the
system's site characterization capabilities.  The  fourth and final demonstration  was a return  visit to
Hanscom AFB with the  intent to perform a more detailed characterization of the site.  The last three field
demonstrations were successful in that the instrument was operational and depth - encoded LIF data were
collected for a significant number of surface push locations.

The matrix-formatted field data are displayed as three-dimensional fingerprints of contamination at given
push locations and depths; they are also reduced in dimension by summing over one or both wavelength
axes and plotting  summed fluorescence vs. depth to facilitate visualization of the approximate extent of the
contaminant plume.   The summed fluorescence  data have been photon - normalized to the  incident
excitation energy  during the push in which they were measured.  Selected EEMs from each push  location
are examined for characteristic patterns and are  qualitatively  compared  to  standard EEMs  of pure
compounds and fuels.

       Several hardware and software challenges remain unmet  and the subject of ongoing research.
                                                VI

-------
                                         CONTENTS
Notice	ii
Foreword	iii
Preface	iv
Abstract	vi
Figures	x
Tables	xii
Acknowledgment	xiii

1.   INTRODUCTION	1
2.   CONCLUSIONS	4
3.   OVERVIEW	7
4.   INSTRUMENTATION	9
A)  Design and Construction	9
    1)  Excitation source and launch system	9
    2)  Fiber cable assembly	11
    3)  Probe design	13
    4)  Scattered light rejection	16
    5)  Detection system	19
    6)  Automatic data acquisition	20
    7)  CPT vehicle installation of instrument	21
    8)  LIF-EEMHPLC	22
B) Instrument Characterization	22
    1)  Wavelength calibration	22
    2)  Calibration curves	23
        a)   Aqueous phenol	24
        b)   JP5 in cyclohexane	25
    3)  Quinine sulfate fluorescence - system diagnostic tool and calibrant	27
    4)  Incident energy vs. quinine sulfate fluorescence calibration	28
    5)  Complex mixture measurements	32
        Jet fuels	32
    6)  Single component EEMs	33
    7)  Multi-component EEMs	34
        Benzene, naphthalene, anthracene	34
    8)  Solid containing samples	35
                                              VII

-------
5.   FIELD WORK	37
A. Methods	37
    1)  Operations	37
        a)  Installation and optimization	37
        b)  CPT-LIF measurements	37
    2)  Data analysis	38
        a)  Accomplishments to date	38
B.  Field Sites	39
    1)  Hill Air Force Base	39
        a)  Site Operations	39
        b)  Field Results	41
    2)  Hanscom Air Force Base I	42
        a)  Site Operations	45
        b)  Field Results	46
            1.   Calibration data	46
            2.   Depth vs. summed fluorescence	47
            3.   Representative EEMs	49
            4.   Peak fluorescence EEMs	51
            5.   Depth corrected EEMs	53
    3)  Coast Guard Station, Elizabeth City, NC	55
        a)  Site Operations	55
        b)  Field Results	57
            1   Calibration data	59
            2   Depth vs. summed fluorescence	61
            3   Summed fluorescence site map	62
            4   Summed fluorescence comparison with laboratory data	64
            5.   Peak fluorescence EEMs	67
    4)  Hanscom Air Force Base II	70
        a)  Site Operations	70
        b)  Field Results	71
            1.   Calibration data	71
            2.   Depth vs. summed fluorescence	73
            3.   Peak fluorescence EEMs	74
            4.   Monitoring well sample	76
                                              Vlll

-------
6.   SUMMARY	78
    A.  Goals Accomplished	78
    B.  Detailed Summary of Goals Accomplished in this Work	78

References	83
Bibliography	86
Glossary	87
Index	89
Appendices	91
    A.  Hanscom I field data
    B.  USCG, Elizabeth City, North Carolina field data
    C.  Comparison of LIF-CPT USCG, NC data with laboratory core sample analysis
    D.  Hanscom II field data
    E.  Improved two fiber probe for in situ spectroscopic measurement
                                              IX

-------
                                           FIGURES
Number
1.   Diagram of the optical layout of the excitation source and
    fiber-optic launch stages	10
2.   Fiber-optic cables for cone penetrometer system	12
3.   Probe inner (a) and outer (b) bodies, window and optical plugs	13
4.   Two fiber probe designs and the fluorescence of 0.1 M phenol solution	15
5.   Fiber holder (b) and probe inner body mounting hardware (a)	16
6.   Ray tracing of 266 nm light in a fiber optic connector with the ends separated
    by a 1 mm gap. Two gap mediums are illustrated: fused silica and air	18
7.   Detection system: fiber optic plug, spectrograph and CCD camera	19
8.   System for automatic data acquisition	21
9.   Standard mercury lamp spectrum used for calibration of the
    spectrograph and CCD detection system	23
10. Calibration curve for aqueous phenol measured with fiber optic LIF-EEM probe	24
11. Calibration curves at each excitation wavelength for JP-5 in cyclohexane mixed with
    sand from USCG station, EC, NC measured with fiber optic LIF-EEM probe	26
12. Fluorescence spectra of standard calibrant, 0.2 g/L quinine sulfate in ethanol,
    at eight of the ten excitation wavelengths normally used	28
13. Calibration curves of incident excitation energy vs. quinine sulfate fluorescence	31
14. EEMs of a) 11 ppm JP-4  in cyclohexane, b)  105 g/L JP-5 in cyclohexane mixed with sand from
    USCG, EC, NC and, c) neat JP-8 mixed with silty clay from Hanscom AFB
    	32
15. EEMs of a) 10 mM phenol in water, b) 0.126 g/L naphthalene in cyclohexane and, c) 50 mM
    anthracene in cyclohexane
    	33
16. EEM of 3% Benzene, 86  ppm naphthalene, 86 ppm anthracene mixture	34
17. Fluorescence spectrum (266.0 nm excitation) from 3 % benzene,
    86 ppm naphthalene, 86 ppm anthracene mixture EEM	35
18. a) Aqueous phenol (940 ppm), b) Aqueous phenol (940 ppm) mixed with Ottawa sand	36
19. Jet fuel contaminated site at Hill Air Force Base situated near SLC, Utah	40
20. Scattered light vs. depth for different excitation channels for push T-13 at Hill AFB	42
21. Map of Hanscom AFB, MA and surrounding area	43
22. Map of Hanscom AFB Former Fuels Area	44
23. Map of CPT-LIF push sites at Hanscom AFB	45
24. Quinine sulfate summed fluorescence averages as a function of push number during

-------
    the field test at Hanscom AFB  	47
25. Summed fluorescence vs. depth for push site 6 with peak EEMs	50
26. Summed fluorescence vs. depth for push site 11 with peak EEM	50
27. Peak EEMs measured in situ at Hanscom AFB, Bedford, MA	52
28. EEMs from the peak fluorescence depth regions of push locations, #6, 7, and 11 	53
29. Map showing the location of the U.S.C.G. support center in Elizabeth City,
    North Carolina, and surrounding areas	55
30. Detailed map of entire U.S.C.G. support center in Elizabeth City, NC, with the area of CPT
    investigation, the fuel farm shown in the upper left hand corner	56
31. Fuel Farm area with the locations of the CPT-LIF measurements (solid circles), and the location
    of drilling sites where core samples were taken for conventional analysis (open triangles)	58
32. Excitation energy as a function of push number during the field test at the U.S.C.G. Support
    Center at Elizabeth City, NC	60
33. Normalized summed fluorescence data plotted with site coordinates to produce a 3-dimensional
    map of the fuel farm at U.S.C.G. Support Station at Elizabeth City, North Carolina	63
34. Depth corrected EEMs for each push location at U.S.C.G.  Support Center, Elizabeth City, NC 69
35. Map of CPT-LIF push locations and site features at Hanscom AFB	70
36. Excitation energy as a function of push number during the field test at
    Hanscom AFB in Bedford, MA	72
37. Depth corrected EEMs measured at Hanscom AFB, Bedford, MA	75
38. EEM of a sample from monitoring well # 9  at Hanscom AFB,
    measured on a medium grade sand	76
3 9. Histogram of laser stability for North Carolina and Hanscom II field operations	80
                                               XI

-------
                                            TABLES
Number
1.   Raman snifter output and excitation energy available	11
2.   Transmission of filters and fiber at each excitation wavelength	17
3.   JP-5 calibration curve regression summary	27
4.   Quinine sulfate fluorescence vs. Incident excitation energy calibration
    curve regression summary	32
5.   Summary of normalized peak fluorescence and depths	48
6.   Summed fluorescence approximate peak depths and contamination depth ranges
    at USCG support center in Elizabeth City, NC	61
7.   Summary of LIF-CPT push locations versus core sample locations for lab analysis	65
8.   Summed fluorescence approximate peak depths and contamination depth ranges
    during Hanscom II	74
                                               XII

-------
                                    ACKNOWLEDGEMENTS

There are many people who deserve credit for their effort at different stages of the project.  I would like to
thank my Project Manager, Bob Lien, for his help and enthusiasm. This was the first project he managed,
and he deserves much credit for his patience, flexibility, and general helpfulness.  Bob has a wonderful
attitude toward science: he loves what he's doing, believes in a hands-on approach, and is always the first
one to dive in and make something happen. Anyone who has seen him driving the 30-ton Mack from one
push location to another, or swing the rods in the hydraulic compartment, or make  his way through  the
lunch buffet at the Great Wall Restaurant, knows what I'm talking about.  Postdoctoral researcher Todd A.
Taylor was involved in the development of earlier versions of our instrument, and helped conceive this
project and write the proposal.  The work was taken over by postdoctoral researcher Jie Lin and graduate
student Sean J. Hart,   who together successfully met the  challenges  of design and assembly of  the
instrument under the heavy pressure of a fiscal year  deadline for the first field test.  When Jie Lin left to
pursue his own academic research and teaching career, Sean provided the continuity and leadership needed
while lie's successor was searched for, hired, and trained. Yu-min Chen arrived near the end of the project
and was an active contributor in the field testing, and the development of several protocols, hardware, and
software  for data acquisition and processing. Jim Bodah of the Chemistry department  electronics shop and
Frank  Mellon  of the department machine  shop each deserve  credit  for  their  input  into  our work.
Throughout the field work, our project manager, Bob Lien, researcher Susan Mravik,  and technician, John
Hoggatt,  were indispensable in the operation of the CPT vehicle and handling of the probe during  pushes.
Dr. John Wilson is acknowledged for coordinating and assisting our work at the U.S.  Coast Guard Station
in Elizabeth City, NC. At Hanscom AFB, the site manager, Tom Best, provided us with much support and
direction during the work in addition to detailed site information and well samples.
                                               Xlll

-------
I. INTRODUCTION

     The possibility of performing in situ (or even simply on-site) chemical analysis of hazardous waste
sites is growing increasingly attractive to government and private parties alike, because of the possibility
of obviating all or some of the following problems: worker exposure, sample collection, labeling, waiting,
unwanted chemical transformations of samples, transport,  waiting, loss  of volatiles,  chain of custody,
waiting, standard laboratory analysis, waiting, etc.  A particularly attractive stage for in situ sampling is
provided by cone penetrometer technology (CPT), which uses a hydraulic system to push a  small-diameter
rod directly into the ground, using the weight of the truck which houses and transports the unit as ballast.
This technology has been used for decades for physical characterization of soils and site geology; recently,
it has become the darling of those wishing to perform chemical characterizations as well.  The technology
appears to  be ideal for insertion of fiber optics which connect the subsurface  sample  to  a  spectrometer
located in the vehicle, at last making good on the promise of fiber optic promoters to provide cheaper
access to the subsurface than the conventional method of well digging. Coupling these optical fibers to
laser light  sources permits  sensitive in situ measurements to be made in many cases.  One  such  case is
detection of fluorescent  analytes,  such as benzene, toluene, ethylbenzene, and xylenes (BTEX)  and
polycyclic aromatic hydrocarbons (PAH).  Several years ago, we demonstrated the feasibility of in  situ
detection of groundwater contaminants using short-wavelength ultraviolet laser light and optical fibers1"4.
Shortly thereafter, we  developed instrumentation for in situ detection of fluorescent analytes by LIF, using
a single wavelength (266 nm) from a compact YAG laser and a simple two-fiber probe which could be
lowered into a conventional PVC-lined monitoring well5.  After brief field-testing of the instrument5, we
returned to the laboratory to  develop more sophisticated instrumentation with multiple-wavelength laser
excitation6"8.  Meanwhile, two other research groups, those of Lieberman9"15 and Gillispie16"21,  pioneered
the  CPT deployment of LIF optical fiber probes.

     The work described in this report has grown out of our laboratory development project mentioned
above.   The fundamental work on the development of the laser /  Raman shifter excitation source  was
paramount in the progress of this project.22  The instrument described herein represents the fourth
generation  of fiber optic LIF-EEM systems developed by this laboratory. However, this work is the  first
that has attempted to field deploy the instrument in a CPT vehicle.

     The field deployment of such a system in a CPT vehicle has several complications  associated with it.
The most important aspect is the mating of fibers  and CPT probe.  The fibers have to be packaged for
their protection during the  CPT push routine, and hardware must be developed to couple the light from
the  fiber to the soil through a sapphire window on the probe  body.  In addition,  the instrument must be
rugged enough to withstand the  rigors  of travel in the rear compartment of the CPT truck.  In keeping
                                                1

-------
with that goal, the system components have been chosen with attention to their individual track records in
field use  in  addition to their scientific performance specifications.  Once the instrument  has been
designed,  assembled,  and installed, the characterization and refinement of its functions  must begin.23
This is the stage of greatest importance, when the instrument's capabilities and shortcomings are revealed
in full.  The complete characterization and calibration of the instrument allow a more robust analysis of
field data to be accomplished.  Towards that goal, several calibrations of the instrument's performance
have been undertaken.

     The potential of our LIF-EEM system for site characterization lies in the type and quantity of data
that are generated in  a single detector exposure.  Fluorescence depends on a number of molecular and
instrumental parameters; if its intensity  dependence on a sufficient number of appropriate parameters is
measured, one can hope to perform a mathematical separation on the data set that replaces the physical
separation of the  various components of the sample that is part of a laboratory analysis.  Thus,  it is a
technique highly  suited to the demands  of "hands-off ' in situ measurement. Of all the parameters that
might be  measured, we decided to measure fluorescence intensity, M, as a function of excitation and
emission wavelengths, AX  and Am, respectively.  The elements, M(^,  AJ,  of an EEM for  a  single
fluorescing compound are given24 by equation 1,
                         , A J = # (A, XA, )«>' ft. Wm X              CD
where k is an instrument-dependent constant, / is the excitation light intensity at the sample, e is the
molar absorptivity, O' is the derivative of the fluorescence spectrum with respect to Am, D is the detector
response, and c is the concentration of the emitting species.  The wavelength dependencies are indicated
in parentheses.  In the  derivation of Equation 1, the optically dilute limit has been assumed, i.e., the
absorbance is  so weak that / is essentially constant across the sample.  For an optically dilute sample
containing multiple noninteracting species, Equation 1 would need to be modified by setting M equal to a
sum of terms of the type shown on the  right-hand side, one  for  each component in the  mixture.
Environmental samples  of interest do not always meet the noninteracting and diluteness requirements for
eq. 1 to be strictly valid. At high analyte concentrations which often prevail in many subsurface regions
of a contaminated site, the quantitative analysis can be hampered by non-linear effects such as the inner
filter (IF) effect and energy transfer (ET).

     Our  instrumentation allows  us to  collect an emission spectrum for each  of up to ten excitation
wavelengths; from these data an  EEM  can be constructed  and mathematically analyzed.  Two of the
multidimensional fitting techniques that may be used are least squares and rank annihilation.  Using

-------
these, the identity and concentration of species contributing to an experimental EEM may be determined.
The approach in the least squares calculation is to express the mixture EEM as a linear combination of
library EEMs for individual fluorescent analytes; the coefficient of each library EEM is simply related to
the concentration of that component in the mixture. The coefficients are arrived at by minimization of an
error matrix equal to the difference between the measured and calculated EEMs.  This method should only
be used when all components of the sample are also in the library.  By contrast, the rank annihilation
method allows the concentration of a given component in the library to be estimated even if fluorescent
species  not in the  library are contributing to the  total  signal.  In this method,  the  component EEM is
multiplied by a scalar and that product is then subtracted from the experimental EEM.  If the scalar is
equal to the amount of that component in the sample divided by the amount in the standard, the resultant
product matrix will have a rank (i.e., number of independent rows and/or columns) one lower than the
rank of the experimental EEM.

     Although these data analysis methods  have been  successful in laboratory analyses of fluorescent
mixtures, there is no guarantee of equal success in the  case of field data.   This is  because  of the
assumption of optical diluteness, the limited number of laser wavelengths that can reasonably be launched
into a fiber bundle, the generally unknown perturbation that the soil and other subsurface components can
produce in the EEM of a given analyte, and the unknown number and type of reference EEMs that are
needed.  One challenge  of this project  (and presumably of future  related work) is to  determine the
applicability of these analysis  techniques to field data. Our expectation is that chemicals can be classified
into groups, and amounts in each group determined semiquantitatively, using the current instrumentation.
If this expectation is realized, it would represent a significant improvement over fluorescence techniques
that produce a single number,  e.g., total fluorescence intensity,  versus depth of the  probe,  and an
important milestone along the way to more complete chemical speciation.

-------
II. CONCLUSIONS

     Our goals of assembling, optimizing and demonstrating a sophisticated CPT LIF probe have been
significantly advanced. In very general terms, the hardware part of the problem has been largely solved,
but the data base, the strategies for data analysis and the generation of software capable of performing the
analyses in a timely manner are very much works in progress. In our opinion, the success of the hardware
development program and the qualitative assessment of the field data provide justification for devoting the
necessary time and energy to the remaining problems.  We have already begun to do so in this and other
related projects.  (See section 6 of this report, Summary, for further relevant discussion in support of the
conclusions presented here.)

     The instrument has been designed, developed, refined, characterized and finally field tested on four
separate occasions at three different  sites: Hill AFB, Hanscom AFB I,  Elizabeth City COS, and Hanscom
AFB  II.   Our research  has succeeded  in  producing  an instrument  capable  of characterizing fuel-
contaminated sites with the measurement of multidimensional data.   We have increased the quantity of
data collected by at least 8 fold (8 data channels compared with 1 for standard LIC-CPT systems). Having
multiple excitation wavelengths, in addition to emission wavelengths, i.e. EEMs, permits one to identify
similarities between the spectral signatures of the field contamination  and reference contaminants,  which
may be  cataloged as individual chemical species, mixtures, or classes of compounds.  This capability has
been demonstrated in a qualitative way for both laboratory and field measurements.

     The instrument has  been used  at the same site, Hanscom AFB,  at different times,  3 months apart.
The site investigated was undergoing large scale remediation at the time and on our return visit we were
able to delineate changes in the site, thought to be attributable to the site cleanup.  This work was valuable
in characterizing  the site  dynamics,  and was accomplished with only  3  days of active CPT sampling on
each occasion.

     Calibration curves have been measured, testing the system's response to different analytes. A study
of an aqueous solution  of phenol, an  important single-ring aromatic  compound, using  266.0 nm
excitation,  indicated a reasonably low detection limit of 2.7  ppm.  Measurements  of single components
and mixtures in various media include the following: aqueous phenol  in solution and on a medium sand;
naphthalene  and anthracene  in cyclohexane; a  benzene,  naphthalene  and anthracene  mixture in
cyclohexane; JP-4 in cyclohexane; JP-5  in cyclohexane and on a medium sand; and JP-8  mixed on a silty
clay. The data on these standard compounds serve as reference tools for the analysis of the field-measured
EEMs.

-------
     An important step for quality  assurance  of fluorescence data is to normalize  the  intensity of
fluorescence to that of the excitation source, so that changes or fluctuations in laser power might not be
misinterpreted  as  changes in contaminant concentrations.  We measured the probe's  response  to a
standard quinine sulfate solution before and after each push; these readings were interpreted using the
results of the calibration of the quinine sulfate fluorescence response to incident energy variations.  This
procedure allowed  the summed fluorescence data to be corrected for excitation energy variation from push
to push, and provided a way to monitor overall instrument performance regularly during field work.

     The field data were analyzed in a standard manner for each field test, starting with an examination
of the quinine  sulfate calibration data.   During the first field test at Hanscom AFB, only before-push
calibrations were performed, and during the subsequent  two, both before and after measurements were
made.  For all field tests, depth vs. summed fluorescence  plots  were prepared for each excitation
wavelength at each push location.  EEMs from the peak fluorescence depth region at each push location
were generated and the patterns compared with those of the reference EEMs. There were several features
in the EEMs measured during each field test that were comparable to some of the reference EEMs. Many
of the in situ EEMs are directly comparable to the  standard jet fuels,  JP-4 and JP-5, confirming the
presence of jet fuel contamination.  There are several field measured EEMs that contain fluorescence
patterns  similar to naphthalene.   In a particular EEM, measured during the first  field test  at Hanscom
AFB, there was contamination with features indicating the presence of anthracene in addition to other
species.

     The success of the EEM analysis, either by visual pattern recognition or mathematical  algorithms,
depends  on the quality and number of the reference EEMs  available for comparison.   A complete
identification of the fluorescent chemical composition of these in situ measured EEMs requires a large
library of  reference EEMs, analogous  to  reference  spectra  in the fields of UV-VIS,  IR, and NMR
spectroscopies,  etc. The uses  of such a library include not only visual pattern recognition,  but also the
mathematical fitting of the library EEMs to the field data.  A relatively small library of reference EEMs
has limited our attempts at visual pattern recognition. This is due  in part to the number of field  tests
performed in the last year of the project.

     Even  with the currently available EEM library,  more quantitative pattern recognition methods are
potentially applicable. These include least-squares, rank annihilation matrix methods, and neural network
pattern recognition.  These  techniques are being investigated  to determine which method will be most
suitable for the  analysis  of field data.   Some of the methods require (or at least are normally implemented
under the assumption of) optical diluteness, which simplifies the mathematical  relationships  involved in

-------
EEM analysis considerably.  It is clear that in the field, this assumption often does not hold, and analysis
schemes will have to accommodate this possibility if even semiquantitative results are to be obtained.

     Our recommendation is that the project be given additional support to permit the completion and
testing of the hardware and the further development of data analysis schemes.  These include stabilization
of the laser output at all desired wavelengths, automation of data acquisition, and reduction of optical fiber
damage. In our estimation,  the hardware issues should be resolvable in one year or less, while the data
analysis project will require about two years, although intermediate analysis  methods can be developed
faster.  At least one or two more field demonstrations, supported by extensive laboratory measurements,
should be included in the development plan.

-------
III. OVERVIEW

     The objectives of this project have been of considerable scope and depth.  While there has been one
main goal, the development of the LIF-EEM instrument for CPT deployment, there have been many
smaller supporting sub-projects.  During the course  of developing the main system, our goals  were
modified, as more was understood about the nature and relative importance of the tasks at hand.  Some of
the sub-goals were accomplished, while others were deemed unfeasible within the scope of the current
project. For example, if work on a sub-task would draw too many resources away from the main  goal,
then it was deemed appropriate to sacrifice it for the overall good of the project.

     In addition to the development and construction of the LIF-EEM instrument, many sub-tasks were
outlined in the original proposal.   These tasks  and goals  are arranged below in  order of decreasing
importance: a) In situ LIF-CPT measurements to be  carried out at a well characterized site, b) Software
development for data analysis, c) The development of an HPLC LIF-EEM system that is field portable, d)
Lamp-based EEM system developed and tested in  CPT vehicle, e)  A bioremediated  site  revisited to
determine short term changes, f)  A video  imaging system for  the  characterization  of soil  types, g)
Incorporation of novel sensors into "spare channels" of our system.

     In retrospect, the tasks and goals described above  may  have been  overly ambitious, given our
inexperience with cone penetrometers and the amount of time and funding available.  Therefore, they
were modified according to  their overall impact upon the  project's main objective.  The field portable
HPLC LIF-EEM instrument was  considered an important and valuable  addition to the LIF probe
instrument.  Its development was begun, but the rigorous field test schedule required all  efforts to be
focused upon the readying of the LIF probe instrument.   The video imaging system, while  a useful
addition, was deemed unfeasible given  current camera technology (soon afterward, Lieberman's group26
proved otherwise). Incorporation of novel third party sensors into  our system has not been attempted, but
the system is sufficiently modular that this could still be done fairly easily in the future.  The installation
of a  lamp-based EEM system was not attempted, although this  possibility was briefly examined in a
related project with separate funding.  Software and algorithms  for data analysis were developed and
written in C++ and were used in the analysis of field data. Our early LIF-CPT measurements were carried
out at well characterized sites.  The final version  of the instrument was tested at Hanscom AFB in
Massachusetts.  This site is  currently under remediation and the extent of the contamination plume is
fairly well defined.  The short-term changes at  a bioremediation site were to be  monitored using our
system. A site where bioremediation was being induced externally was not available, but Hanscom AFB
is  undergoing  remediation including soil vapor extraction  (SVE), groundwater treatment facilities, and
significant natural bioremediation as evidenced by the large amounts of bacteria that are routinely found
                                               7

-------
in the treatment  hardware.   A  return visit was made 3 months after the first characterization and
significant changes were discovered.

     The main purpose of this project, the development and field deployment of our LIF-EEM instrument,
has been accomplished. The instrument has been developed and has successfully completed three field
tests at two different sites. Two  of the field test data sets (Hanscom AFB) were measured three months
apart, enabling us to determine  that the remediation efforts at that site have produced changes in the
contaminant levels.  An HPLC system was procured for use in the field and the design of the LIF-EEM
HPLC detection system has been furthered.  Data analysis and processing programs have been developed
that enable the data reduction and presentation of field data to be accomplished in a short amount of time.
In addition, the design of an automatic data acquisition and storage system  has been completed.  This
system allows the automatic collection of depth data and synchronized LIF data collection  and storage
with real time display of total summed fluorescence for each excitation channel.  This automated data
collection feature has been constructed and tested in the CPT vehicle and laboratory, but has not been used
during a field test.

-------
IV. INSTRUMENTATION

A)  Design and Construction

1)  Excitation source and launch system

     The excitation source in our LIF system is comprised of a Q-switched flashlamp-pumped Nd:YAG
laser (Surelite I, Continuum, Santa Clara, CA) operated at a frequency of 20 Hz.  The laser has frequency-
doubling and quadrupling crystals installed, converting its 1064 nm fundamental beam to 532 nm and 266
nm radiation respectively. The 266 nm UV radiation serves as a pump beam for the Raman shifter.  The
Raman shifter is a high pressure gas cell with UV transparent windows, filled with methane and hydrogen
(in a ratio of 55:45). When the 266 nm pump beam is focused into the Raman shifter, a SRS (stimulated
Raman scattering) process occurs.  This process generates many beams22 (> 20) of different wavelengths
that can then be separated and used.  Most of the SRS beams are of longer wavelength than the pump
beam and are called  Stokes-shifted beams.   There are, however,  several beams that  are  of  shorter
wavelength than the pump, and these are called anti-Stokes shifted beams.

     A diagram of the excitation  source and the launch system is  shown in Figure 1.  Two dichroic
mirrors (CVI,  Livermore, CA) are used to remove the fundamental (1064 nm) and the second harmonic
(532 nm) beams from the laser output. The fourth harmonic (266 nm) is reflected by a mirror, focused by
a lens (f.l. = 75 cm), and reflected by another mirror through a 75 cm long Raman shifter.23'25  The beam
exiting the Raman shifter is collimated and passes through a prism system  for the  re-direction and
dispersion of the beams  on the optical breadboard (TMC, Peabody, MA).  The Raman shifted beams of
different wavelengths pass through a prism system where they  are  dispersed according to wavelength.
Five prisms are employed, four of them having a refracting angle of 60° and the other having a refracting
angle of 45°. The incident angle on the first prism and the angles between prisms are adjusted  so that the
aberration for the central beam is minimized.  The beams are refracted approximately 180° on the  optical
breadboard.  The last two prisms are again 60° but have larger  apertures compared to the first three to
accommodate the dispersing beams. The dispersed beams emergent from the prism system are focused by
a lens (f.l. = 75 cm) and launched  into optical fibers to be delivered to the sample.  Each fiber is held on
an X-Y positioner (Newport, Irvine, CA), mounted on a breadboard post, to allow adjustment of the  fiber
launch end.  For each mount, the fiber is held in a fiber-optic chuck (FPJ-H, Newport, Irvine, CA) that is
mounted on the X-Y positioner with provision for the fiber  end  position to be adjusted along  the Z-axis
(optical axis) as well.

-------
                                                 Fiber optic
                                                 Launch Stages
                                                                             Dichroic mirror
                                                                             Dichroic mirror
                                                                    Fiber-optics
Figure 1. Diagram of the optical layout of the excitation source and fiber-optic launch stages.

     Up to ten (eight used in this work) Raman shifted beams were selected based on their pulse energies
and the wavelength intervals (i.e., approximately equal wavelength  spacing).  The wavelengths were
calibrated with a mercury lamp and the pulse energies were determined with RjP 735 and 765 probes and
an Rj 7000 energy meter from Laser Precision (Utica, NY).  The wavelengths and pulse energies for the
ten beams are listed in Table 1.  The  total dispersion distance  for these ten beams at the focal plane is
about 10 cm.  The pulse energies of the beams range from 3.8 to 848 |jj.  Some of the beams  are too
intense and therefore damage the fiber at the focal point. For our system, which  was not fully optimized
in this regard, we found that beams which have pulse  energy  higher than 100  |jj were likely to cause
damage to the fiber, particularly around the focal point of the beam. Other factors which influence optical
fiber launch  damage  include the light wavelength, fiber  diameter  and surface condition.   The most
practical solution to the fiber damage problem was to  move the  fibers away from the focal points.  A
position was found for each fiber where  an appropriate power density (which did not necessarily
correspond to the entire beam) could be launched into the fiber without damaging it.

     The beams were launched into a 10 m fiber cable which  was attached to a 20 m cable at a panel
using standard ST bulkhead connectors.  The energies emergent from the other end of the 30 m cable at
the sample were measured for each of the beams. These excitation energies are listed in the last column
of Table 1.  They are dependent on the amount of light launched into the fibers, the  attenuation of the
beams by the fibers, and  the coupling efficiencies of the connectors.  As  can be seen, the  beams at
intermediate wavelengths  have the highest excitation energies  available.  This  is because those beams
have high total energies and medium attenuation by the fiber optics.  The beam at 266 nm has the  highest
                                               10

-------
total energy, but only a small portion of the beam is launched into the fiber (to avoid fiber damage) and
the attenuation by fiber at this wavelength is very high. As a result, the excitation energy available at 266
nm is not very high.

Table 1. Raman shifter output and excitation energy available.
Wavelength
(nm)
257.5
266.0
278.5
288.4
299.1
314.9
327.7
341.6
362.1
378.8
Beam origin3
(H, M)
1,-1
0, 0
l,-2
0, -1
-1, 0
0, -2
-1,-1
-2, 0
-3, 1
-1, -2
total energyb
(|jj/pulse)
9.6
848
9.8
98
197
188
92
138
7.4
3.8
excitation energyb
(|jj/pulse)
0.01
6.4
0.04
15
65
40
10
25
0.03
0.02
aH, M are the numbers of hydrogen and methane quanta shifted, respectively.
b Total energy is that of the laser beam before launch; excitation energy is that available at the distal end
of the fiber for fluorescence excitation.

2)  Fiber cable assembly
     The choice of optical fiber was based upon several competing factors.  These include mechanical
strength, resistance to light damage during launch, light attenuation, and cost.  The mechanical strength
is very dependent upon the  jacket, or protective coating, used to cover the fused silica fiber core and
cladding.  The  diameter of the  fibers is  also very important.  The fiber diameter in part determines the
minimum fiber  bend radius which the fibers can withstand without damage or loss of performance.  The
larger the core diameter, the  larger the minimum bend radius.  This is an important consideration because
the fibers make  a tight turn between rods in the racks of the CPT vehicle. In addition, the diameter has an
effect on the ability of the polished fiber launch face to withstand damage.  In general  it seems that the
larger diameter core fibers have the greater resistance to light launch damage.  This may be because
damage occurs at the core-cladding interface when the beam is larger than the fiber core diameter.
     The fiber was manufactured by Polymicro Technologies, Inc. (Phoenix, AZ).  The diameters for the
silica core  (doped with fluoride), silica cladding,  and polyimide coating are  400,  440,  and 470  mm,

                                                11

-------
respectively. The fiber cables were assembled in-house.  The cable tubing is composed of a polypropylene

inner tube (i.d.=  1.2 mm), a PVC outer tube (o.d.= 3.8 mm), and a layer of Kevlar yarn in between.  Up to

four fibers can be pulled through the cable by using a UV-curing epoxy to glue the fibers onto a nylon

thread inside the inner tube.


     Before the  fibers were  polished or installed into the connectors, one half to one centimeter  of

polyimide coating was removed by burning, and the exposed portion of the fiber was cleaned with ethanol.

The fiber ends  were polished with  12,  3, 1,  and 0.3 |j,m lapping papers and  the finished ends were

examined under a microscope for surface smoothness and absence of fiber edge chips or lips.
                         system
                        10 launching fibers
                         1 spare fiber
                              20 fiber-optic
                               connectors
                          10 excitation fibers
                          10 receiving fibers
                           1 spare fiber
   4ji   to detection
3j  J   system
10 detection fibers
 1 spare fiber
                                                        between optical
                                                        system and
                                                        hydraulic room
 fiber-optic cables
 with 10 channels,
 each of which has
 an excitation fiber
 & a detection fiber
                                                        threaded through
                                                        cone penetrometer
                                                        rods and connected
                                                        to optical probe
cross section view of
fiber-optic cables.  The
numbers in the circles
are numbers of fibers
in each of the cables.
                        Figure 2. Fiber-optic cables for Cone penetrometer system.



     The configuration of the fiber-optic cables is shown in Figure 2.  A short cable (length = 10 m) runs

between the optical system and the hydraulic room in the cone penetrometer vehicle.  This cable splits into

two branches, each containing ten fibers, at the  optical bench: one is connected to the launching stages

and the other is interfaced to the detection system.  A long cable (length = 20 m) containing all 20 fibers
                                                  12

-------
goes through the cone penetrometer rods and is connected to the optical probe.  The two cables are
connected in the hydraulic room by 20 fiber-optic connectors (ST type,  Augat, Kent, WA).  Basically,
there are 10 channels available, each of which has  a pair of fibers, one for delivering the  light to the
sample and other for collecting the fluorescence.  The lengths of fiber used in this  work are short relative
to those used by other more established systems which are capable of pushing to 100 ft to 150 ft.15'21  We
chose to use shorter fibers to enhance the possibility of BTEX detection using the shorter uv wavelengths
available from our system. Attenuation of laser power becomes significant at wavelengths less than about
290 nm.
3)  Probe design
     The diagram for the entire probe is  shown in Figure 3. The outer probe body is a heat-treated,
hardened steel tube that is threaded on both ends to permit attachment to the CPT push rods.
                                                       Outer body
                         inner body
                                   hole for
                                   cable
                                         1.5"
                                    optical
                                    plug
                                            window
                                             plug
                                      o-ring
                        Figure 3. Probe inner (a) and outer (b) bodies, window
                                        and optical plugs.
The CPT rod screws into the top of the probe outer body, while the cone tip is screwed onto the bottom of
the probe body. The probe outer body has ten holes bored through it in a column that are countersunk to
                                               13

-------
accommodate the sapphire window holders.  The probe inner body fits into the probe outer body and is
sealed by o-rings at the top and bottom.  The inner body is a half cylinder with 10 holes bored along its
length with counter-sunk holes to accommodate the optical mating between the sapphire window and the
fibers. The inner probe body is held in place by the sapphire window holders that screw through the probe
outer body. The fiber-optic cable is pushed through the hole on the top of the probe body and sealed with
RTV silicone sealant.  The fibers are installed into the optical plugs that are held onto the probe body by
set-screws. The inner probe body is inserted into the outer housing and they are aligned. The o-rings on
the top and bottom of the probe body provide sealing so that the groundwater does not leak into the probe.
Then the  window plugs are  screwed into the probe  body, holding the probe body  and outer housing
together. The space between the probe outer body and the sapphire window are sealed using both copper
washers and RTV silicone sealant.

     The hardware for the first optical probe,  used  during the first field test at Hill Air Force  Base, was
machined by Applied Research Associates, Inc. (South Royalton, VT) as per our design.  The  optical
design involved a mirror and a short focal length lens mounted in a cylinder with a hole and set screw for
the optical fibers to be mounted and interfaced with the mirror / lens system.25  The light from the
excitation fiber is reflected by the first surface reflecting mirror, then focused by the  lens onto the outer
surface of the sapphire window, where the window contacts with the soils.  Similarly,  the fluorescence of
pollutants is collected by the detection fiber.  The two fibers  (excitation and detection) are held inside a
stainless steel tubing.  A semi-circle filter is placed in front of the detection fiber to cut  down reflected and
scattered excitation light.  For different excitation beams, different filters were used.

     In practice several factors contribute to inefficiency in  this optical system.   Firstly the lens  and
mirror were not anti-reflection coated (due to cost and time constraints); this led to light reflection losses
that  approach the unacceptable limit.  When assembling the  lens and mirror, the tiny size of the  optics
made the insertion and cementing of them in the  cylinder difficult, with smearing of the cement and some
misalignment of both lens and mirror. For these reasons, an alternative that would address these  issues
was sought after the first field test.

     The next optical probe design involved a much simpler approach: have the optical fibers bend 90°
and make  optical contact with the sapphire window. This method of light delivery (parallel fibers against
a sapphire window) has been used by others17.  We incorporated a fiber end geometry that maximized the
overlap  of the fiber optic distal cones at the outer surface of the sapphire window.  Two geometries were
compared with the standard parallel fiber model and  one was chosen for its simplicity, ruggedness  and
good performance.  Figure 4  is a design and fluorescence spectrum comparison of the three geometries
fabricated,  tested and compared: parallel, beveled ends,  and angled flat ends.  It can be seen that the
                                                14

-------
fluorescence for either of the new probe designs, A and B in Figure 4, is larger than that for the simple
two  parallel  fiber design.  Using the modified  two fiber probe  designs,  a  detailed  study of probe
improvement factors (ratio of new probe to old design:  B/A and C/A) was undertaken.  Improvement
factors up to  14 were seen for the new probes in different solid-containing media.27  Refer to appendix E
for the details of the two fiber probe experiments.
                         Intensity
                         X1000
                         (counts)
                                   200   240   280   320
                                        Wavelength (n
                                                     360  400
                      Figure 4. Two fiber probe designs and the fluorescence of
                      0.1 M phenol solution. A) parallel fibers, B) beveled ends,
                      and C) angled fibers with flat tips.
     The use of the new probe design required a system to push the fibers against the sapphire window.
The major problem in implementing this with the probe inner and outer bodies was that when the fibers
are extended the  inner probe body cannot be put inside the outer probe housing.  The solution to this
problem involves holding the fibers in  the probe inner body with a  set-screw in the  sapphire  window
threaded opening until the assembly is completed. With the fibers held back, the probe inner body can be
inserted into the  probe  outer body.  Then the set-screw is removed  and the fibers in their holder are
released and the sapphire window can be screwed into place with the fibers pressing against it.  The final
                                                15

-------
design of the probe plug and the fiber holder is shown in Figure 5.  The probe plug is a cylinder with a
center hole and another underneath.  The center hole is for the fiber holder and the one underneath is for a
key attached to the holder. The key keeps the fiber holder in position and prevents rotation.  A spring in
the lower hole is used to push the fiber holder forward. The fiber holder slides forward and
         Figure 5. Fiber holder (b) and probe inner body mounting hardware (a).
backward in the probe plug.  The preparation of the fibers involves the epoxying of the fibers at the set
angle 16.4°.  The fibers, exposed from the main fiber cable, are covered with PVC tubing and slipped into
the fiber holder tube.  Then epoxy is inserted into the assembly to set the fibers against the holder walls.
A small bead is allowed to form on the tip so that the epoxy can be polished away exposing the fiber tips.
4)  Scattered light rejection

     In a fluorescence experiment, some of the excitation light is coupled towards the detector through
either Rayleigh scattering or reflection off optical elements and soils used in the measurement.  If the
detector is sensitive to this exciting light, it may become saturated with this residual scattered light. Often
it is necessary to remove this light due to its spectral interference with the desired fluorescence response.
The most common way to accomplish this light filtering is to place a long-pass absorption filter in front of
the detector.  Ideally, using this technique, the scattered light is reduced or removed and the fluorescence
is not affected in any way.  There are only certain cut-off wavelength filters commercially available, and
                                                16

-------
these are not always of the ideal cut-off wavelength for the Raman shifted beams used.  In some cases,
some fluorescence must be sacrificed for the sake of scattered light removal.

     For different excitation beams, filters with different cut-off wavelengths were selected, based on the
transmission of filter and fiber (30m long). It was intended not to completely remove the reflected and
scattered light, so that small  residual Rayleigh scattering and water Raman  signals could be used for
diagnostic purposes (potentially, to obtain information about the sizes of soil particles or water content, or
as calibration wavelengths in the detection system).    The wavelength of water Raman shifted light is
calculated  from  the excitation wavelength and  a water Raman  shift frequency of 3400 cm"1.  The
transmission of the filter and fiber at each of the wavelengths is determined  from their spectra.  Total
transmission, calculated as the product of transmission by filter and fiber, is listed in Table 2.  As seen,
more than  90 %  of the reflected and Rayleigh scattered light will  be removed by the filters (except the
longest wavelength for which an appropriate filter is not available).
Table 2. Transmission of filters and fiber at each excitation wavelength."
Wavelength
(nm)
257.5
266.0
278.5
288.4
299.1
314.9
327.7
341.6
362.1
378.8
Schott
Filter Number
(Hill AFB)
WG280
WG280
WG295
WG305
WG320
WG335
WG345
WG360
GG395
GG395
Schott
Filter Number
(All other work)
None
WG280
WG295
WG320
WG345
WG345
WG360
WG360
GG395
GG395
Total transmission (%)
WLRaylelRh WLRaman
0.4
4.3
2.9
9.0
5.4
2.3
3.0
3.6
7.9
36
18
31
36
45
53
58
63
69
68
76
       1 WLRayleigh is the wavelength for Rayleigh scattering and reflected light,
        and WLRaman is the wavelength for water Raman shifted light.
       Normally, the filter is placed directly before the detector's sensing surface, where the light is being
delivered to the detector.  In a system such as ours where there are multiple channels containing different
information (i.e.  different exciting wavelengths), there is generally  a  different  filter for each channel.
                                                17

-------
Since each channel is defined by the emission fiber carrying fluorescence to the detector, there must be a
filter not much larger than the fiber diameter, 400  |jm.  With the fibers arranged in a column at the
entrance slit of our spectrograph (see below) with at most a 200 |jm spacer in between them, the filters
would need to be cut into 1A mm strips and aligned with respect to the appropriate fiber.  It is a challenge,
first, to cut the filters cleanly, and, secondly, to align them  accurately.   Furthermore, if the channels
needed to be changed, (as occasionally happens during field work, due to fiber damage) the filter attached
to one fiber on the detection end would be inappropriate for the new channel.

       The solution adopted involves the use of the standard fiber optic connectors used in our system to
join the excitation source and detector to the optical probe in the hydraulic room.  The approach involves
the placement of filters in the fiber optic connectors.  The filter material is cut and filed into  circles that
closely fit in the fiber optic adapter that mates  the  two connector ends.   The connector  ends then are
inserted and locked into place in the adapter, but are separated by the thickness of the  filter.  Naturally,
there is a loss associated with the separation of the fiber ends but it is comparable to the loss incurred for a
standard fiber connection.

       The loss depends upon the material separating the fiber connectors and the distance of separation.
The larger the separation, the larger the fraction of light that is not coupled into the receiving fiber. In
addition,  the refractive  index  of the material between the fiber ends affects the light refraction and
therefore the loss. Figure 6 is a ray tracing illustration of 266 nm light in a fiber optic connector for two
              12.7
                                                Gap
        - Gap=Air
• Gap=Fused silica
                                                                               Fiber-optic
              12.7°
                                                                                   12.7°
Figure 6.  Ray tracing of 266 nm light in a fiber optic connector with the ends separated by a 1 mm gap.  Two
gap mediums are illustrated: fused silica and air.
                                                 18

-------
different cases: a fused silica flat and an air gap. The filter ray trace is not shown because its refractive
index at 266 nm (n = 1.52) is very close to that of fused silica (n = 1.54). The ray tracing was done for the
limiting ray emerging from or entering the fiber, as calculated from the fiber's numerical aperture: N.A. =
0.22 = n sin 9, where n is the refractive index and 9 is the angle of the light ray with respect to the fiber
axis.  It can be seen from the ray traces that the air gap should have a larger loss associated with it due to
the larger angle exiting the fiber and thus a larger fraction of the beam's cross section not being coupled
into the receiving fiber.

     The advantages  of filters include their being easily manufactured,  implemented,  and replaced  if
needed, with an acceptable light loss.   The inclusion of filters in fiber optic connectors is consistent with
the desire to keep the system as modular as possible:  If a channel must be changed,  then its filter can
easily be exchanged with a different one for the new channel.

5)  Detection system

     The detection system is composed  of an  imaging spectrograph (SpectraPro  150,  Acton Research
Corporation, Acton, MA) and a CCD detector (TE/CCD-576EMUV from Princeton Instruments, Trenton,
NJ).
                                                19

-------
                                    shutter
Spectrograph
                       CCD
                      detector
                 ..concave
                  mirror
                                                                  .concave
                                                                   mirror
                                                entrance slit
                                                fiber-optic plug
                           Fiber-Optic Plug

1 0 fibers
O O
0 0
                                             interfaced to spectrograph
                    Figure 7. Detection system: fiber optic plug, spectrograph and CCD camera.

Ten detection fibers are held in a column by an aluminum plug, which is interfaced to the spectrograph.
The grating is set so that wavelengths in the region of 250 to 500 nm are detected in first order by the
CCD detector. The fiber optic interface plug, the spectrograph and CCD are shown in Figure 7. The light
path through the spectrograph is shown starting with the light from the fibers entering the spectrograph's
entrance slits.  The  image  of the fibers is  focused onto  the  CCD chip held at the  exit plane of the
spectrograph. A shutter between the CCD and spectrograph controls the CCD chip exposure time.

     The CCD  detector has a UV coated EEV chip with  576 by 368 pixels, which is cooled to  -30  °C
thermoelectrically. The long axis of the CCD chip is set horizontally and used as the wavelength axis.
For this work, the CCD was binned into eight channels to detect the fluorescence from the eight detection
fibers.  Each channel is composed of 18 - 20 rows of pixels with five to ten unused rows in between to
reduce the cross-talk between the channels due to charge blooming and spectrograph aberrations.  Along
the long x axis there  are 576 pixels defining the maximum  emission resolution. Our actual resolution was
2.0 nm - 4.0 nm depending upon the  spectrometer  slit  width.  We  could  have binned pixels  in this
direction to correspond to the actual resolution, but this was not done during data collection. The software
used to collect  spectroscopic data was CSMA for DOS (Princeton Instruments,  Trenton, NJ) for  work
                                               20

-------
through the first field test and Winspec for Windows 3.1 (Princeton Instruments, Trenton, NJ) for all
work thereafter.

6)  Automatic data acquisition

     In the field it is desirable to have the ability to collect both LIF and push depth data simultaneously
with little operator intervention.  The system, during the first 4 field tests, had no automation hardware or
software; therefore the probe was stopped at a given depth and operators manually initiated data collection
and recorded the tip depth.  After these initial experiences, a system to enable automatic data acquisition
was developed.

     The automatic data acquisition system depends  upon  the use of two computers:  one for LIF data
acquisition and storage (high speed Pentium PC), and the other (486 PC) as a slave for the collection of
depth data and triggering of the first computer.  The  scheme for automatic data acquisition is shown in
Figure 8.
         LIF Data pc
                                   Network connection
                                                         Depth / Trigger pc
                                                          ADC Board
             Keyboard
                                                                 Connection Box
                                                          CPT Depth
                                                          Transducer
       Figure 8. System for automatic acquisition of LIF and depth data.
     The depth transducer in the hydraulic room of the CPT vehicle feeds a signal to a connection box,
which serves to route the signal to the appropriate places.  The voltage from the depth transducer is
                                             21

-------
captured by an 8 bit ADC board, and custom software written in visual basic (VB) converts the voltage
signal to a depth reading. At depth intervals (0.125 to 1 foot) set in the VB software, two trigger signals
are sent to the LIF data PC's keyboard by the depth / trigger PC to simulate commands.  The first trigger
starts data collection, saves the binary data file, and creates the new file for the next trigger.  The second
trigger initiates the conversion of the binary files to ASCII and the summation of the LIF data in each
strip. The summed data can be transferred to the depth / trigger PC for graphical display or can be plotted
on the data PC directly.  The software developed for data acquisition and post processing also has the
ability  to  display in real time the  progress of the  push:  depth  vs. total summed fluorescence  for each
channel. This system was not developed in time to be used during any of the field work described.

7)  CPT vehicle installation of instrument

     The fiber optic LIF-EEM instrument must be installed in the rear room of the  CPT vehicle.  All  of
the equipment must be secured and protected for the journeys between sites  and for maneuvers on site.
Towards that end,  our system was ruggedized: the optical breadboard was mounted on a table built from
Unistrut metal bars that were mounted on spring shock absorbing mounts (VMC, Bloomingdale, NJ). The
electronic  components were placed in standard 19" rack mount cases and secured in a rack that was bolted
to the vehicle floor.

     The laser system and the data acquisition equipment are kept in the rear room while the probe and
fiber connectors are kept in the front hydraulic room. A short cable (length = 10 or 5 m) runs between the
optical system and the hydraulic room in the cone penetrometer vehicle.  Further details of the fiber optic
cabling have been described in section 2) above.

8)  LIF-EEM HPLC

     The  development and incorporation of HPLC into the field work was planned in two stages:  1)
Incorporation  of standard HPLC in the field using rapid  extraction techniques  for  quick screening
methods and, 2) The  development of the LIF-EEM HPLC system for the analysis of core samples for
comparison with in situ measurements. The standard HPLC system has been purchased, assembled, and
tested.  The methodology for rapid extraction and  sample pre-treatment before HPLC sample injection has
been chosen but not tested.  The LIF-EEM detection system was designed to be added to the  standard
HPLC  system so that the standard and LIF-EEM  detectors operate in series.  It involves the use of fused
silica capillaries (Polymicro Technologies, Phoenix, AZ) to carry the post column eluent. The fused silica
capillary has its jacket removed enabling the light from the excitation fibers to be focused into it and the
fluorescence to be collected by detection fibers oriented 90° from the excitation fibers. Due to a rigorous

                                               22

-------
field testing schedule for the in-situ LIF-EEM probe (three field tests in four months), the development of
the HPLC capability was not completed during the project.

B.  Instrument Characterization

1)  Wavelength calibration

     One of  the  first  and most  important  calibrations needed for the  collection  of useful  spectral
information is that of wavelength.  The CCD camera can rotate within the spectrograph's exit opening
before tightening  a set-screw to fix its position.  The rotational alignment must be carefully checked by
observing the fiber  images on the monitor to  ensure that the  wavelength axis of the  spectrograph is
parallel to the long axis of the CCD.  Once this is accomplished, the CCD is then carefully focused to
bring the CCD chip surface into the focal plane of the spectrograph.  Once both alignments have been
accomplished, the position is set and calibration can begin. The  usual output of the CCD  data acquisition
is pixel numbers (the light intensity at each pixel across the horizontal axis).  To calibrate  to wavelength a
standard mercury vapor lamp is used.
                              Intensity vs. Pixel Number
              70000 j

              60000 - -

              50000 - -
           g  40000 - -
          •55  30000 - -
           it
          £  20000 - -
              10000 --

                 0
JL
                          100    200     300    400     500
                                       Pixel Number
                                                                600
                                                                       700
   Figure 9.  Standard mercury lamp spectrum used for calibration of the spectrograph and CCD
   detection system. The 6 peaks are assigned, in order from left to right: 253.65 nm, 296.73, 312.57,
   365.02, 404.66, 435.83.
                                               23

-------
This is positioned a 1A m away from a fiber bundle (a 5 m section of fiber cable with a spectrograph plug
on one end and fiber connectors on the other). This is connected to the short cable in the CPT vehicle and
the connectors attached to the emission fiber  channels. The result is that mercury light is delivered to the
spectrograph in exactly the same way (through the same fiber in the same positions) that calibrant and
contaminant fluorescence will be delivered. Figure 9 is a typical mercury lamp spectrum showing the five
largest lines in the wavelength region 250 nm - 450 nm.  Using the known wavelengths of the major lines
in the mercury spectrum, the data can be easily converted from pixel  numbers to wavelengths using a
linear or cubic polynomial fit.

2) Calibration curves

    The overall response of the system to fluorescent compounds depends  upon many factors beginning
with the excitation source.  The more excitation light that  can be  delivered to the sample (up to the
compound's saturation or decomposition point), the more fluorescence will be seen.  The amount of 266
nm light available from the laser to pump the Raman shifter will directly affect the amount of Raman
shifted light available to be launched into the fiber optics.  The position and condition of the fiber at the
launch point has  a  dramatic effect on the amount of light delivered.  The fiber optics have significant
attenuation in the deep  UV; therefore, when delivering short wavelength excitation  beams they will
deliver less light to the sample.  Other factors affecting the overall sensitivity of the instrument include the
overlap of the fiber optic distal  cones on the sample surface,  the efficiency of the fiber optic connectors,
the coupling efficiency of the fiber optic / spectrograph plug,  the throughput of the spectrograph, and the
sensitivity of the  CCD chip pixels.   Overall performance has been quantified by generating calibration
curves for a specific compound, and for a fuel in solvent on sand.

a)  Aqueous phenol

     In Figure 10, a plot of peak fluorescence response vs. aqueous phenol concentration is shown.  The
linear fit of the data is enhanced considerably when the highest concentration is eliminated, suggesting
that the upper limit of phenol's signal response  linearity is  about 100  ppm.  By extrapolation with the
minimum detection signal set to baseline counts plus 3  times the baseline standard deviation, the LOD
(limit of detection) for phenol in water was determined to be 2.7 ppm.
                                               24

-------
Normalized Fl
3r\T7_i_A'3 _, 	
.Uii+UJ
2^T7_l_A'3
. Jii+UJ
2AT7J^A'3
.UE+U3
1CT7 1 AO
.5E+UJ
1 AT^-LAQ
l.Uii+UJ
5AT7_I_AO
.Uii+U2
A AT"1 1 AA
I
_ . - 1


/-"'"
/

U.UJzrrUU 1 1 1 1
O.OE+00 l.OE+02 2.0E+02 3.0E+02 4.0E+02 5.0E+02
Aqueous Phenol concentration (ppm)
  Figure 10. Calibration curve for aqueous phenol measured with fiber optic LIF-EEM probe with excitation
  at 266 nm.  The equation of the line for the entire data (dashed line) set is y = 5.18x + 125.44, with r2 =
  0.9820, while the equation of the line for the data set less the highest concentration (solid line) y = 8.89x +
  6.03, with r2 = 1.0000.
b)  JP5 in cyclohexane

     The calibration of a LIF-CPT instrument using a complex mixture, known to be similar to the type of
contamination found in situ, is a common method of converting LIF signal into a quantity well understood
by the site characterization community12.   A common contamination  source at military petroleum-
contaminated waste sites is jet fuel of the JP-X series, whereas at industrial sites diesel, gasoline and fuel
oils are a more common source of contamination.  Commonly, sites with older spills have either aviation-
gas (av-gas) or JP-4 contamination. Currently, fuels  such as JP-5 and JP-8 are being used and the former
fuels are difficult to obtain. The compositions of these fuels are known to be similar, but  with differing
concentrations  of fluorescent species28.  The response of the instrument when the probe is exposed to a
fuel is due to fluorescence from the many different fluorescent components present in jet fuels, potentially
including interactions among the component species.

     The  use of a generic standard to quantify the  in situ contamination, while  useful, has many
difficulties. Ideally, the in situ contamination should very closely approximate the fuel standard to allow a
                                                25

-------
positive identification to be made.  This may sometimes be the case, but even if the original contaminant
material were  identical to  the  standard,  weathering  of the  contaminant (e.g.,  loss  of volatiles,
bioremediation) generally occurs at different rates for the different components, resulting in an imperfect
match between the weathered contaminant and the standard.  However, the calibration of the instrument
using a  standard fuel is important for the characterization of the instrument even if the resulting
calibration data are not applied to the actual field data.

     In Figure 11, the response of the instrument to JP-5 in cyclohexane mixed with a medium grade sand
is shown, measured on-site.  Excess JP-5 solution was mixed with the sand and the mixture was allowed
to stand.  The excess solution was removed prior to measurement in a trough style sample holder onto
which the probe is placed with the sapphire windows against the  sample.  The concentrations of JP-5 in
cyclohexane were chosen to allow us to determine the  response of the system to both low and high
concentrations of jet fuel. The concentrations were in the 0.6 % (w/w) to 72.2 % (w/w) range. The lower
concentration is similar to a smear zone or lightly contaminated zone, whereas the upper concentration is
more similar to a pure product region, or saturated  zone; both concentration regimes are expected to be
present at jet fuel contaminated sites.

     The responses of the  instrument to the jet fuel on sand were linear with correlation coefficient
squared values, r2, generally above 0.9, with an average value of 0.951. There may be inner filter effects
hampering the measurement of the upper concentration,  563  g/L. When the fits are  repeated with this
point deleted, regression coefficients  (average  0.965)  and intercept values generally improve.   The
regression coefficients for each excitation wavelength are listed in Table 3.  These data  were produced
from detector dark current subtracted fluorescence sums and normalized with excitation energies derived
from quinine sulfate
                                                26

-------
                                              299.1 nm
                                                                                                341.6 nm
    4*
    u
    §
    w
    Wl
    Si
    o
   '•
   •_
   o
288.4 nm
            4*
            u
                                                          o
                                                          S
266.0 nm
                                                                                                327.7 nm
                Z.
                                              257.5 nm
                                                                                                314.9 nm
            Concentration JP-5 (g/L)
                   Concentration JP-5 (g/L)
Figure 11. Calibration curves at each excitation wavelength for JP-5 in cyclohexane mixed with sand from
USCG station, EC, NC measured with fiber optic LIF-EEM probe.
calibrations before and after each measurement. This is the same procedure used to analyze the field data
(refer to section B.4).  This was done to allow the direct comparison of the in situ summed fluorescence
quantities to  be  correlated with a known amount of jet fuel, thought to be similar to some of the
contamination seen at all of the sites involved.  A comparison of field data EEMs (Figures 26, 27, 33, 36)
with the JP-5  EEM (Figure 14 b ) indicates that there are some in situ EEMs that are directly comparable
to the reference jet fuel  EEM, while others are unrelated indicating an entirely  different type of
contamination.  For the related EEMs, the signal may be attributable to similar jet fuel  contamination
levels.
Table 3. JP-5 calibration curve regression summary.  All data have been normalized using excitation
energy data except the 257.53 nm channel where a quinine sulfate fluorescence sum was used
without conversion.
                                              27

-------
                         Full data set
Less highest concentration
Beam
-1
0
1
2
3
4
5
6
Excitation A,
(nm)
257.5
266.0
278.5
288.4
299.1
314.9
327.7
341.6
Slope
0.0209
2.044E-08
-
3.988E-08
6.478E-09
4.514E-10
4.343E-10
2.577E-11
y-int
0.0714
1 .967E-06
-
4.554E-07
1 .646E-07
2.622E-08
-3.211E-09
1.439E-09
r7
0.9726
0.8551
-
0.9637
0.9804
0.9366
0.9979
0.9515
Slope
0.0339
6.643E-08
-
6.515E-08
1.161E-08
1.104E-09
3.226E-10
5.181E-11
y-int
-0.0509
5.341E-07
-
2.174E-07
4.805E-09
5.894E-09
1.317E-09
6.277E-10
r7
0.9874
0.9745
-
0.9406
0.9990
0.9928
0.9916
0.8706
     In calculating the LOD (limit of detection), a baseline sum y-intercept plus 3 times the error in the
baseline was used.   This corresponds to the summed fluorescence base line  in  the  absence  of  JP-5
fluorescence. The calculated LOD for JP-5 in cyclohexane on a medium sand in 0.29 g/L.  This LOD is
for the channel most sensitive (containing the greatest fluorescence) to  JP-5: the 288.40 nm excitation
channel. All other channels should have higher LODs due to the lower light levels that are closer to the
signal baseline.

3)  Quinine sulfate fluorescence - system diagnostic tool and calibrant

     A  system calibration was performed before each push to accomplish two goals: 1) confirm system
operation and channel integrity and 2) provide a measurement of beam intensity for photon normalization.
The calibration compound used was quinine sulfate, well known as a fluorescence standard and to a lesser
extent as a quantum counter.  This calibrant was dissolved in ethanol and not the usual dilute H2SO4 due
to damaging reaction of the latter with the steel probe body.

     All of the probe channels report fluorescence, although not uniformly. This is due to several factors
including the fiber optic attenuation of the strongest UV beams,  and the low energy of the high order
Stokes and first anti-Stokes beam.  The net result is that the central channels (288.5 nm - 314.5 nm)  have
the largest signal. The emission  spectra at the  excitation wavelengths  measured for standard  quinine
sulfate solution in ethanol (0.2 g/L) are shown in Figure 12.
                                               28

-------
     GO
     O
     O
     GO
     a
     (D
     (D
     O
     GO
     (D
Excitation X (nm)
 257.7
 266.0

 278.4

 288.5
 299.1
 314.5
 327.3

 341.6
                    Emission A
Figure 12.  Fluorescence spectra of standard calibrant, 0.2 g/L quinine sulfate in ethanol, at the excitation
wavelengths normally used.
     For most fluorescent molecules in condensed phases, the distribution of fluorescence emission as a
function of emission wavelength is independent of excitation wavelength; therefore the wavelength of
fluorescence peaks of quinine sulfate at the different excitation wavelengths should be identical. This is
not the case for the spectra shown in Figure 12 where there are apparent shifts in the emission maxima.
The spectral shifts are due to the different cutoff filters used in each channel to reject scattered light.

4)  Incident intensity vs. quinine sulfate fluorescence calibration

     The correction of the probe-measured data for instrumental artifacts must be  accomplished if the
field data are to be compared to  standard reference EEMs. If the reference EEMs are not standard, but are
                                              29

-------
produced on the same instrument, then full correction of the field data is not required, but reference EEMs
for all  analytes of interest must be collected  whenever  the  instrument configuration  changes,  e.g.,
increasing fiber length to access deeper plumes. However, even when reference EEMs have been obtained
on the instrument as configured for the field, variations in excitation energy in each channel can be
appreciable and must be accounted for.

     An expansion of Equation 1 to include the quantitative details of the fluorescence  dependence is
shown in Equation 2:
                                                                       CCD(^) c.        (2)
Comparison with Equation (1) shows that we have expanded I(AJ  into separate factors: Ipi(^x), the pre-
launch intensity of the excitation laser beam,  L("L), the launch efficiency of the beam,  and TX(^J, the
transmission of the excitation fiber.  Similarly,  the detection sensitivity D(km)  has been expanded to
include Tm(km), the transmission of the emission fiber, F(km), the transmission of the selected cutoff filter,
S(^m),  the spectrograph efficiency,  and CCD(^m), the  sensitivity of the detector itself.  Lastly,  the
instrument constant k of Eq. 1 has been replaced  by k 'G, where G is a geometrical factor accounting for
the collection efficiency of the probe.  G and S should have little or no intrinsic dependence on L,, but the
fact that  each excitation channel uses a different pair of  fibers and  a different path through  the
spectrograph can result in different numerical values of these two quantities for each excitation channel.
For example, SfAj can vary for emission fibers positioned at different heights along the entrance slit of
the spectrograph because of vignetting (smaller effective aperture of spectrograph for off-axis points) or
other imaging aberrations.

     For a given instrument configuration, we can reasonably expect the  last four instrument factors in
Eq. 2 to remain constant.  Each of the component factors of I(LJ  can change: Ipi(Lj varies in the short
term because of shot-to-shot variations  in the pump laser power, which are  amplified by the nonlinear
processes  used to generate new frequencies, and it steadily degrades over the long term as the flashlamp
ages and  as optical elements degrade (especially, uv mirrors losing their high reflectivity).   L(LJ may
vary from shot to shot as well; if the beam profile at the launch end of the excitation fiber has a spatial
extent  about the same as or larger than the fiber aperture, then any small variations in  this profile can
change the fraction of the beam that is  successfully launched  into the excitation fiber.  In addition, any
mechanical instabilities that result in movement of either the laser beam or the fiber end will  reduce
L(LJ from its optimal value. Jarring shocks such  as may be experienced when the cone penetrometer hits
a cobble might induce such changes instantaneously during a push.  Finally, TX(LJ can change due to
                                                30

-------
solarization, the loss of transmissivity experienced by a fiber that is strongly illuminated by ultraviolet
light. This effect occurs relatively slowly over time, over hundreds of thousands or millions of laser shots,
but at high laser  repetition  rates,  its effects might  be significant  over the duration of a single  site
investigation.   Of course,  instantaneous fiber damage may occur if allowable peak power levels are
exceeded, but we have already discusses strategies to avoid this more acute case of laser-induced decreases
in Tx(h).  By contrast, the emission fiber is not exposed to focused laser light at all; although the intensity
of scattered laser light incident on the emission fiber  is much higher than that of fluorescence,  the laser
light has been significantly attenuated by the excitation fiber and scattering by the sample, so solarization
of the emission fiber is unlikely, as implied above.  G should remain constant for a given probe channel
unless physical damage occurs; this would be detectable by means of the visual inspection that is done
routinely before and after every push.

     In this work, we compared field EEMs only to reference EEMs generated with the same instrument;
this allowed us to ignore both corrections for D^) and the channel dependence of G.   In the absence of
visual detection of damage,  G is considered constant for each channel.  We  utilized the average of
fluorescence measurements of a standard solution of quinine sulfate in ethanol (into which the probe is
dipped) immediately before and after each push to provide an estimate ofl(^) during that push.  In order
to facilitate cleaning of the probe after this measurement,  as well as for economic reasons, the quinine
sulfate was not used at high quantum-counter type concentrations which would render its  fluorescence
response independent of its own wavelength-dependent absorptivity.  Therefore the fluorescence response
in each channel is not simply proportional to I(^) but contains a dependence on the concentration of
quinine  sulfate  and  its wavelength-dependent  absorptivity.   (This dependence could  be removed,
mathematically, in theory, but in practice the correction is hampered by our non-standard probe geometry,
for which published analyses  of signal level are nonexistent, and the fact that our quinine sulfate is neither
quantum-counter thick nor optically thin.)

     This complication, simply another channel-dependent correction factor, need not concern us as long
as we are comparing EEMs from the same instrument. One simply normalizes the fluorescence intensity
in each  channel by dividing it by the average of before  and after triplicate measurements of quinine sulfate
fluorescence signal for that push; the resulting intensities are counts/quinine sulfate counts, or effectively
dimensionless.  We initially performed all our fluorescence normalizations this way.  However, to better
characterize our system and to pave the way for more complete data reduction schemes in the future, we
also calibrated the quinine  sulfate response in each channel by comparing it  to the laser intensity at the
sample  as  measured by  laser energy  probes.  A  post calibration  (after  field work was complete) was
performed on the instrument  using the same fiber optics, probe assembly, and operating parameters. The
incident excitation energy was varied by adjusting the laser output power, and the fluorescence response of

                                                31

-------
quinine sulfate was  measured as a function of laser  pulse energy for each  channel. The laser  pulse
energies, measured before and after the quinine sulfate response and averaged, were converted to photon
numbers and plotted against quinine sulfate fluorescence counts  as shown in Figure 13. The  data are
plotted with x and y  error  bars, and the  regression line reflects a  weighted  fit  using the  summed
fluorescence error bars.  A summary of the regression coefficients is given in Table 4, r2 values of 0.9 or
higher. The quinine sulfate measurements made during the field test were then converted to the number
of photons per pulse incident on the sample.  This number, obtained from the calibration curve, was then
used for the photon normalization of the data from each push.  The resulting photon normalized number
has units of fluorescence counts per photons of excitation light delivered.  Post calibration data were  taken
for  all excitation channels except 257 nm and 278 nm.  For these channels,  the summed fluorescence
could not be converted to photons per pulse,  so the raw number of summed fluorescence counts was used
as the normalizing factor.
                                                              3.5E+06

                                                              3.0E+06

                                                              2.5E+06

                                                              2.0E+06

                                                              1.5E+06

                                                              1 .OE+06

                                                              5.0E+05

                                                              O.OE+00
                                                                  4.E

                                                              4.0E+06
    341.56 nm
A
   •10    1.E+13    2.E+13    3.E+13
                                                                       1.E+12  2.E+12  3.E+12  4.E+12
                   I.E+09  1.E+10  2.E+10  2.E+10  3.E+10
              Excitation Energy (photons / pulse)
                                                              1.4E+06
                                                              1.2E+06
                                                              1 .OE+06
                                                              8.0E+05
                                                              6.0E+05
                                                              4.0E+05
                                                              2.0E+05
                                                              O.OE+00
                                                                 4.E+11    2.E+12   4.E+12   6.E+12   8.E+12
Excitation Energy (photons / pulse)
Figure 13.  Calibration curves of incident excitation energy vs. quinine sulfate fluorescence.
                                                  32

-------
Table 4. Quinine sulfate fluorescence vs. Incident excitation
energy calibration curve regression summary.
Excitation A (nm)
266.03
288.40
299.10
314.89
327.69
341.56
Slope
4.49E-06
2.04E-06
1.57E-06
1.73E-07
5.76E-07
8.26E-08
y-int
-16233.30
3986.95
-678.89
6523.23
3641.85
3632.82
f
0.8708
0.9830
0.9928
0.9929
0.9473
0.9081
5)  Complex mixture measurements
Jet fuels
     Lab measurements of jet fuels were made in various media.  JP-4 was measured in cyclohexane (11
ppm), JP-5 measured in cyclohexane mixed with a medium sand (105 g/L), and JP-8 was measured neat
on a soil sample (silty clay) from Hanscom AFB.  The EEMs of these mixtures are shown in Figure 14,
a,b,c. In Figure 14 a, the fluorescence intensity is low due to the low concentration of fuel relative to the
others.  In both Figures  14 b  and  c,  the fluorescent intensities are much higher, as expected, due to the
higher fuel concentrations.

     The  JP-4 EEM has significant  signal  arising from both the 288.4 nm and 299.1 nm excitation
channels, with a peak EEM fluorescence at (A*, Am) of 299.1 nm, 340 nm.  Similarly, the JP-5 EEM has
signal arising from both the 288.4  nm  and  299.1  nm excitation channels, but  with a peak  EEM
fluorescence at (Ax, Am) of
                                               33

-------
                                                                     IZ
Figure 14. EEMs of a) 11 ppm JP-4 in cyclohexane, b) 105 g/L JP-5 in cyclohexane mixed with sand from
USCG, EC, NC and, c) neat JP-8 mixed with silty clay from Hanscom AFB .  In each case, the vertical axis
represents excitation wavelength (nm) and the horizontal axis represents emission wavelength (nm).
288.4 nm, 330 nm.   The JP-4 EEM has other features, notably  the long wavelength fluorescence
extending from 375 nm to 450 nm in the  341.1 nm excitation channel.  The JP-8 sample has a different
EEM pattern characterized by a peak EEM fluorescence at (A*, Am)  of 266.0  nm, 340 nm.   There is
additional
fluorescence at excitation wavelengths 288.4 nm - 327.7 nm.
6)  Single component EEMs

     The  collection of single component EEMs with the LIF probe is important for  EEM analyses.
Having a  library of carefully measured EEMs  of suspected contaminants is essential  for a complete
comparison of unknown field  data  EEMs with the standards.  In some cases  the fitting of standard
mixtures such as jet fuels (JP - 4, 5, 8) is acceptable, but in other situations a more specific answer is
sought. Towards that end EEMs of single compounds were measured in differing solvents.

     Figure  15   shows EEMs  of a)  10 mM  phenol in distilled water, b) 0.126 g/L  naphthalene  in
cyclohexane, and c) 50 mM anthracene in cyclohexane.  As seen, EEMs  provide very fine fingerprints of
the compounds which can be used for identification and quantitation of fluorescent contaminants.  These
three EEMs illustrate distinct
                                               34

-------
Figure 15. EEMs of a) 10 mM phenol in water, b) 0.126 g/L naphthalene in cyclohexane and, c) 50 mM
anthracene in cyclohexane. In each case, the vertical axis represents excitation wavelength (nm) and the
horizontal axis represents emission wavelength (nm).
regions of compound fluorescence excitation and emission that our system is capable of measuring.  In
addition, these compounds represent three major classes of aromatic contaminants: single,  double and
triple ring species.  The EEM of the phenol sample has its peak EEM fluorescence at (A*, Am) at short
wavelengths:  (266 nm, 300 nm), while the naphthalene EEM is shifted towards the center of the EEM
with maxima at (289 nm, 325 nm). Furthermore, as the aromatic conjugation is increased with increasing
ring number, the excitation and emission are further  shifted  to  longer wavelengths such as  with
anthracene, Figure 15, c, whose EEM maximum occurs at (315 nm, 400 nm).

     The naphthalene EEM, Figure 15b, and JP-4 and JP-5 EEMs, Figures 14 a  and b, are directly
comparable. One would expect the reference samples (naphthalene, JP-4 and JP-5) to resemble each other
because of the presence of naphthalene fluorescence in the jet fuel EEMs.  This is  consistent with our
knowledge that the characteristic fluorescence signatures of these jet fuels are dominated by naphthalene
and  its derivatives,  which have higher absorbances and fluorescence quantum yields than the BTEX
components.

7)  Multicomponent EEMs
Benzene, naphthalene, anthracene
     An important measurement made was a 3 component mixture of representative  1, 2, and 3 ring
compounds.  This measurement allows us to clearly examine  the instrumental response from three
important classes of compounds.  The standard mixture measured contained: 3 % (v/v) benzene, 86 ppm
(w/w)

-------
       £
       =
           30000
           25000
           20000
            15000
            10000
             5000
 Figure 16. EEM of 3% Benzene, 86 ppm naphthalene, 86 ppm anthracene mixture in cyclohexane.
 FL intensity is given in counts.
naphthalene and 86 ppm (w/w) anthracene in cyclohexane.  The EEM of the  3-component mixture is
shown in Figure 16. The characteristic quadruplet of anthracene dominates, and to shorter excitation and
emission wavelengths the naphthalene doublet (330 nm peak) and finally the emission of benzene (280
nm peak) can be seen in the 266.0 nm excitation channel.   The single  spectrum for  the 266.0 nm
excitation channel for the standard mixture is shown in Figure  17.  While the anthracene fluorescence in
the long wavelength excitation channels of the EEM in Figure 17 is  still the most intense, the smaller
naphthalene and benzene signals can clearly be seen. There are three distinct regions of fluorescence
emission in this EEM, and the excitation channels contain little overlap.  The 266  nm channel has
emission from all three compounds, and the  299  nm channel has both naphthalene and anthracene, but
the rest remain separated.  This separation  shows  the ability to discriminate between these classes of
compounds in the emission dimension. The excitation dimension is omitted only to more clearly show the
emission spectrum detail.
                                               36

-------
                     2000

             aj  a
             u  5
             S  o
             Si
             o
             3
                       220.0
                               270.0
                                       320.0
                                               370.0
                                              (nm)
                                                       420.0
                                                               470.0
           Figure 17. Fluorescence spectrum (266.0 nm excitation) from 3 % benzene, 86 ppm
           naphthalene, 86 ppm anthracene mixture EEM.
8)  Solid-containing samples

     Aqueous phenol solution was measured in Ottawa Sand (EM Science, Gibbstown, NJ).  The EEM of
phenol in solution alone is shown in Figure 18a for comparison, and the EEM of phenol in Ottawa sand is
shown in Figure 18b.  The absolute peak signal in  solution was 860 counts vs.  554 counts when the
phenol solution was mixed with sand. This 36 % loss  of signal is at least partly due to the excitation light
penetration depth being reduced in Ottawa sand (particle size 420-595 mm) vs. that in solution.27
                                               37

-------
      vo.
Figure 18. a) Aqueous phenol (940 ppm), b) Aqueous phenol (940 ppm) mixed with Ottawa sand.
                                              38

-------
V. FIELD WORK

A. Methods

1)  Operations

a)  Installation and optimization

    The first task when arriving at a site for field testing is the physical installation of the system in the
rear room  of the CPT vehicle.  This involves the mounting of the optical breadboard on the spring
mounted table, and all  electronics in the  rack mount cabinet.  All other instruments must be secured
before movement of the CPT.  The fiber cable ends are threaded through the hole in the wall separating
the instrument and hydraulic rooms.  The two cables are connected at the fiber connector box and the
cable ends are attached to the appropriate instruments.  The probe end  with  the ten fiber holders is
threaded through the push rods and then the assembly of the probe can commence.

     After the physical installation has been completed, the calibration and optimization of the system can
begin. Wavelength calibration is done first, as described above, before the filters are placed in the fiber
optic connectors.  The filters are then installed and the operation of the  system is evaluated beginning
with the 4th harmonic output power of the laser and the appearance of the Raman shifted beams.  The
beams are launched into the fibers and the identities  of the  beams at the probe end are confirmed.  The
probe can then be placed in the quinine sulfate standardization solution for optimization. The x, y, and z
positions of each fiber launch stage are adjusted until a maximum signal is seen due to the increased
amount of excitation light being launched.  After obtaining acceptable signal levels in all channels, the
system is ready for performing field measurements.

b)  CPT-LIF measurements

     Each push  begins with a measurement of the standardization solution:  quinine sulfate in ethanol.
This measurement performs two functions: 1) It alerts the operator of any  changes in the performance of
the system and,  2) It records the specific performance data for  the given  push  that is used in the  data
analysis routines to power normalize the resulting push data.  The push begins at low resolution (1/2 to 1
foot) until contaminant signal is seen, when the push is slowed and the resolution increased (1/4-1/8 foot).
Another mode of operation is possible where the  CPT  operator advances the probe at a standard rate
(ASTM, 2cm/s for example)  and  the LIF instrument operator collects data continuously.   If any push
locations appear to have more regions of interest missed during the descent, they can  be probed on the
                                               39

-------
way up, although this type of analysis may be plagued by soil and contamination smearing against the
sapphire window. At the end of the push, the probe is cleaned and the same quinine sulfate solution is re-
measured to document any changes in the instrument performance.

2)  Data analysis

a)  Accomplishments to date

     A system such as our LIF EEM probe generates a large quantity of data.  Several data analyses can
be performed at different levels of complexity for various purposes.   The simplest analysis scheme we
have used involves total fluorescence in each channel, i.e., the summation of detector pixel intensities at
all wavelengths greater than the excitation wavelength of the channel in question.  This amounts to data
reduction where the emission wavelength information is suppressed in order to allow rapid visualization
of total signal as a function of depth.  This type of data analysis has been utilized to produce the plots of
total signal vs. depth similar to total or peak fluorescence  intensity vs. depth provided by other LIF-CPT
tools (in addition to their other display modes). The main difference is that our system has 7-10 signal
vs. depth plots (7 - 10 excitation X) for each push location compared with one for other LIF-CPT systems.
This simple analysis has been incorporated into our automatic data acquisition and processing software. It
provides a crude estimate of contamination level versus push depth.  The differences among the signal vs.
depth plots for each excitation wavelength provide evidence of changes in  contaminant composition
(and/or soil type, moisture content, etc.).

     An algorithm and program have been developed for a more robust version of the above data analysis.
The simple data analysis described above does not correct for any instrument functions and is therefore
only a crude indicator of the contamination levels encountered by the probe.  The robust version includes a
background (in  this context, detector background, due  to dark  counts and  read  noise)  subtraction,
summation after excitation peaks, normalization, and full error propagation of the resulting intensities.
The normalization amounts to the division of the summed intensities by a number  specific to each
channel. This number is the average of the background-subtracted summed fluorescence of the quinine
sulfate measured before and  after each push. This procedure  accounts for  variations in the system
performance from push to push.

     A further variation on this theme is the use of a calibration curve, fluorescence of quinine sulfate as a
function of incident  excitation energy, for photon normalization as discussed in  detail above (II.B.4).
With this working curve, and the relation Ephoton = hc/A*, the quinine sulfate sums can be converted to the
number of photons per pulse for each push. The data are  treated in exactly the  same way, but with the
                                                40

-------
normalization data in photons per pulse being calculated from the working curve.  It must be noted that
while the normalization procedure accounts for push to push  variations, short-term variations (point to
point within a push) cannot be corrected.

     An extension of the summed fluorescence analysis to obtain a concentration plot for each push
location has been done.  This involves the use of a system calibration curve with a known standard fuel
mixture. The standard fuel fluorescence is treated in exactly the same way as the field data.  Using this
calibration curve, the field data can be converted to equivalents of the standard  calibration fuel and the
concentration can be plotted as  a function of depth.   The standard fuel  chosen must have significant
spectral similarities with the EEMs seen at the site for this to be valid.

    Further compression of the  multidimensional data to give a single "grand" fluorescence sum  as a
function of depth might be desirable to provide operators with a real-time display, or for general use as a
simple indicator  of  contaminant distribution.  Because variations in laser  output, both absolute and
relative, in the various excitation channels, such a summation should use background-subtracted, photon-
normalized summed fluorescence from each channel as input into the "grand sum."  Because the different
excitation wavelengths of our current probe sample the soil at different depths,  some interpolation or other
approximation must  be made to obtain the "grand  sum" at a specified depth.   Work is in progress to
provide such plots for each push at Hanscom I and II and EC.

B.  Field Sites

1)  Hill Air Force Base

     The site chosen for the first deployment of our LIF fiber-optic probe sensor was Hill Air Force Base,
near Salt Lake City, Utah.  It was selected by EPA personnel at RSKERL (Robert S. Kerr Environmental
Research Laboratory) in Ada, Oklahoma due to their previous experience at that site. A map of the AFB
is shown  in  Figure  19 with  the area of investigation circled.  The area of  CPT investigation was an
abandoned fire fighting pit and fuel storage facility.

a)  Site Operations

     The field work performed at Hill  Air Force Base (AFB) in Salt Lake City, Utah on September  12-20,
1994, represents the  first CPT vehicle deployment of our system25'29.  The system deployed was the first
version that had just  been assembled prior to departure for the field exercise. The field test was preceded
by an installation in late August, 1994 of the system in the CPT vehicle stationed at the RSKERL in Ada,
                                                41

-------
Oklahoma. The CPT truck with the installed system was driven from Ada, OK to the AFB in Utah where
work began September 12, 1994.  This first field exercise was hampered by the arrival of the designed
probe components only days before departing for the field test. In fact several of the needed items were
shipped from the probe manufacturer (ARA, South Royalton, VT)  directly to the field location.   The
installation of the probe took three days and on September 15, 1994, the first push was made.
                                     OPEBABLi UNIT 1
                                LOCATION Of OPERABLE UNITS
                                       mourn i.a
Figure 19. Jet fuel contaminated site at Hill Air Force Base
situated near SLC, Utah. The circle indicates the location on the
base of the LIF-CPT testing.

     Two of ten sapphire windows were broken during the first push, presumably because some of the
window plugs could not be screwed into the probe body completely and, thus, the sapphire windows were
not flush with the probe surface.  They were  replaced and other window plugs were readjusted.  No
window breakage occurred during the following two pushes.  About fifty exposures of the CCD camera
detector were taken at different depths during each of the pushes (to a depth of about 20 feet).  Each
exposure acquired nine fluorescence spectra each of which was excited by a different laser wavelength.
                                               42

-------
(Two of twenty-one fibers in the long cable were broken and, thus, only nine channels were available with
the remaining nineteen fibers).

     Three pushes were completed in total during the field test that was cut short by damage to our laser
power supply, probably caused by inadequate generator capacity.   The CPT vehicle generator was not
capable of supporting both our laser and the rod steam cleaning system simultaneously.  Nor could the
instrument remain powered up when the vehicle was in motion, maneuvering between push locations at
the site.   This necessitated turning the laser  system on and off  many  times during the day, a very
unfavorable situation. On one particular occasion, the electrical system was shut down and then restarted
without warning.   The  generator  surge pulse damaged  a PC logic board in the laser  resulting in
unrecoverable failure. The three pushes accomplished were later found to be in a region that had  < 200
ppb contamination (equivalents of fuel), i.e., below detection limits as measured by another more
established LIF system.  In the absence of fluorescence data, the scattered light was analyzed as a function
of depth.

b)  Field Results

     As discussed above, the scattered light (reflection from surfaces, scattering by soil particles,  and
Rayleigh scattering) is reduced by long-pass cut-off filters.  More than 90% of the back-scattered light is
removed by the filters and by attenuation over the optical fibers. It  is intended  not to  completely remove
the scattered  light  because (1) complete removal  results in significant  reduction of  the  fluorescence
emission and (2) the small residual scattered light can be used for diagnostic purposes.

     Figure 20 shows profiles of scattered  light vs. depth  profiles for different excitation  wavelengths
obtained at Hole T-13 during the field test. The variations in the scattered light appear to be caused by the
changes in soil samples because (1) the variations in the scattered light are larger than the variations in
intensities of the laser  beams and (2)  the  patterns of the variations are fairly consistent among the
different wavelengths.  The profiles seem to  indicate that there were some noticeable changes at depths of
2, 13, and 17 feet.
                                                43

-------
                             257.7 nm
                             0     100
    Intensity
    (counts)
 266.0 nm
 0    300
278.4 nm
0     150    300
                     Depth
                      (ft)
                             288.5 nm
                            0    10000  20000
 299. Inm
, 0    10000
314.5nm
0     1000   2000
                             327.3 nm
                             0    20000
                                              341.6 nm
                   362. Inm
                   0   300   600
                  Figure 20. Scattered light vs. depth for different excitation channels for
                  push T-13 at Hill AFB.
2)  Hanscom Air Force Base I

     Hanscom AFB is located in the central part of Middlesex County, Massachusetts, approximately 14
miles northwest of downtown Boston and 11.5 miles south of downtown Lowell, Massachusetts. The Base
comprises approximately 826 acres and occupies land in the towns of Bedford, Concord, Lexington, and
Lincoln as shown in Figure 21. Prior to 1973, the primary mission of Hanscom AFB was the operational
maintenance of fighter  aircraft and research  and development (R&D)  support.  In August 1974, the
original
lease  permitting the operation and  maintenance of the runway and flightline  activities was canceled
following the termination of Air Force flying activities in 1973.  The airfield reverted to state control and
was renamed L.G. Hanscom Field, currently operated by the Massachusetts Port Authority (Massport) as a
                                                44

-------
civilian airport.  The remainder of the property was retained by the Air Force to support the current
mission which is Research and Development and Command, Control, Communications, and Intelligence.
Figure 21. Map of Hanscom AFB, MA and Surrounding Area.

     The  site  under investigation is  the  former fuels area which included off-loading,  storage and
dispensing facilities  for jet  fuel,  aviation gasoline,  and  more  recently, No. 2 fuel  oil.   The  site is
approximately 7.5 acres in size and is located in the northeast corner of Hanscom AFB,  south of Massport
Runway 29 and the Shawsheen River. The  storage facilities included two 500,000 gallon above ground jet
fuel tanks, five above ground 50,000 gallon aviation gasoline tanks (ASTs)  and six underground  12,000
gallon jet fuel/aviation gasoline tanks (USTs).  Also included at the site were fuel loading/off-loading
stands, a rail siding, and pump houses.  The aviation gas system was taken out of service in 1973 and the
50,000 gallon ASTs were removed in 1986. The 500,000 gallon jet fuel storage tanks were abandoned in
1973; however, they were not removed until 1990 when both tanks were pumped free of oil, cleaned and
dismantled.  During the oil shortage period in the 70's the westernmost 500,000 gallon tank was used for
the storage of No. 2 fuel oil.

     In 1990 Hanscom AFB discovered evidence of historic releases of jet fuel and/or No. 2 fuel  oil.
Investigations  to date confirm that there  are two  separate  plumes  of  light non-aqueous phase liquid
(LNAPL) and that there are multiple locations of petroleum  contaminated soils and BTEX contaminated
groundwater.  The sources of this contamination are believed to be the underground piping associated
with former rail siding unloading systems and one or both  of the former  500,000  gallon above ground jet
fuel storage tanks. A map of the site is shown in Figure 22.
                                               45

-------
     The availability of information regarding location and quantity of contamination  and type (pure
product vs. smear zones) was important in choosing this site.  The history of the site provides us with
additional clues about where the product is traveling to or where it might be pooling and collecting.  In
Figure 21, the
                        )         >     —. —• —~      *•
                      -•^:r~ ' .-^^   i=
Figure 22. Map of Hanscom AFB Former Fuels Area.
Shawsheen River can be seen immediately northwest of the site.  The Shawsheen River recharges the
Town of Bedford wellfield  1.5 miles downstream and the Town of Burlington has a water supply inlet 4
miles downstream.  With the river so nearby,  it is imperative to know the boundaries of the pure fuel
product plumes and the extent of BTEX contaminated groundwater.

     Hanscom AFB  is conducting a CERCLA Remedial  Investigation of the site  to  determine the
complete nature and extent of contamination and has conducted interim removal actions since the original
                                              46

-------
discovery of the petroleum contamination.  Hanscom AFB is currently conducting a CERCLA Removal
Action which includes groundwater treatment and soil vapor extraction (SVE) systems.   The current
system involves nine (9) dual phase recovery wells and three (3) dual phase recovery trenches.  The soil
vapor extraction system is designed to remove volatile compounds from the soil  gas whereas the
groundwater treatment system removes floating product and BTEX.  Over 2,000 gallons of product has
been recovered since this dual phase system commenced operation in October 1995.
                                                  Old jet
                                                  fuel tank
                                                  location
                                            Fuel recovery system
  Old Fuel
  pipeline
          Old railway line
          for fuel delivery
                                    O MW-IC
                                                    O MW-9
Figure 23. Map of CPT-LIF push sites at Hanscom AFB.
a)  Site Operations

     The CPT vehicle arrived at Hanscom AFB  on June 20, 1996, in order for the installation to be
completed before EPA personnel arrived for the start of data collection.  Preparation of the vehicle for the
arrival of the instrument was needed before  any items were brought to the  site.   The moving and
installation of the instrument began immediately thereafter.  The adaptation of the laser electrical supply
cables to the output connectors of the portable generator that had been rented was needed. The extra long
                                               47

-------
insulated cable (30 feet) was constructed and the entire electrical system was tested before the actual test
began.  As soon as all items were in place and in order, the installation of the probe began.  During the
probe installation, the first difficulties were encountered: the inner probe body was scratched and it would
not fit into the outer body housing. The tolerances of the two probe pieces are quite small and any bends
and/or scratches make the fit difficult.  This was overcome  by a trip  to a  machine shop where the
tolerances  on the pieces were opened up slightly.   After  the reworking of the probe inner and outer
housings,  the installation went  smoothly, but we had been put behind schedule by two  days.  The
installation was completed on June 25, when the EPA personnel arrived.  One more  day was necessary for
instrument setup and calibration and by the  morning of June 27, we were ready to begin operation.
During the following  field measurements, 14 locations were pushed in 2.5 days of sampling. The push
locations for Hanscom AFB I are shown in Figure 23.  In brief, in situ fluorescence was detected at push
sites 1-13 and none was seen at push  14.
b)  Field Results23'30

     Our primary objective during this  work was the  demonstration of our instrumentation, not site
characterization.  The outcome of our field test at Hanscom AFB will be evaluated on that basis.  While
the instrument and the analysis of the data are still the subject of much research and development, we note
the successful collection of a significant number of calibration and measurement data sets.  Approximately
30 files were recorded during a push including calibration data.  The total amount of raw data per push
stored on the computer's hard disk is approximately 2 MB; with files generated during analysis that
number is easily tripled. Even in the crudest mode of operation (detect or non-detect at > 20 depths / push
location) the data show the capability of  the  system to rapidly characterize a site and the  extent of
contamination.  The data shown and discussed are by no means a complete and representative sampling of
the type and scope of data analysis that is  possible with the data collected.  In fact the current data analysis
consists of one of the simplest analysis schemes involving raw data reduction.  Future work involves a
more complete mathematical analysis of the data using matrix least squares fitting algorithms with library
EEMs for speciation.

     In the assessment of our instrument's operation during field work, the first step is the analysis of the
calibration data. The next step is the production of plots of summed fluorescence vs. depth for each push,
which allow easy examination of the peak contaminant signals versus depth.  The current analysis does
not involve EEM analysis but qualitative discussion of representative EEMs is presented and demonstrate
the "fingerprint" quality of the EEM data  obtained.
                                                48

-------
1.  Calibration data

    The quinine sulfate solution was measured in triplicate before each push. The results are shown
graphically in Figure 24.  The changes in the fluorescence response  of the calibrant, quinine sulfate,
throughout the entire field test are given for each of the excitation wavelengths.  In addition, error bars
(standard deviations at the 95 % confidence level) for each quinine sulfate summed fluorescence average
are given.  The percent standard deviation ranged from 0.2 to 28.  The data were used to normalize the
fluorescence  data from each push  relative  to the incident excitation intensity available during that
particular push.    The variations  were  large  and could  have  adversely  affected  any  attempts  at
contamination quantification.
(1
5.E+06
4.E+06

2.E+06
1.E+06




2
'a? C
•** 3.E+05T
a
3
g 2.E+05-
x — '
Sj 1.E+05-
g



I
g ' 2
3 b
•Q 5.E+05
1>
S 3.E+05
s
3 1.E+05

r™i



2
a
1.E+05]

8.E+04-
6.E+04'
4.E+04'
2.E+04-





n
2






























1
3




1
3

n




3






n


3































4





4

n




4






h
* rfi



t


rfi 3.E+06'

n n n n -^

i*i


5 6 7 8 9 10 11 12 13 14 23
g




n


4.E+06-
rFi rii J,
* r m 3.E+06-
nO n fin III ::::


ri
0 E+QQ i ' ' ' n"
5 6 7 8 9 10 11 12 13 14 23
f














r*i
















i
4












567


I

n






4



[i




567



1

n M
5 6 7 8 9 10 11 12 13 14 23
e
	 1 5.E+06-T




n


4





4.E+06-
3.E+06-
2.E+06-
nnnnnnnnn



n

n
5 6 7 8 9 10 11 12 13 14 23
Push Number








*
PI













P
a






4567







4


r









567





r
|





3




1


-




8





^

8




n n
9 10 11




n n
9 10 11






Ei







f*





l




1





12
















1
13








f\




14




n H r*i
12




1

9 1011




f








9 1011








|f|




13






12














12



1





14



I*l







13 14



7t|



13




|*|l

|


14
Push Number
Figure 24.  Average quinine sulfate standard solution  summed fluorescence (detector  counts) calibration
summary for each of the excitation wavelengths at each push location made at Hanscom AFB. Insets a - e
represent summed fluorescence averages at excitation wavlengths (nm), 257.7, 266, 278.5, 288.4, 299.1, 314.9,
327.7, and 341.6  respectively.

Depth vs. summed fluorescence data
     During each push, a single measurement (2 seconds exposure, fluorescence from 40 excitation laser
pulses summed) was made at each depth. Each measurement contained 8 channels of data, one for each
                                                49

-------
of the excitation wavelengths used.  The summed fluorescence vs. depth data are shown for pushes 2-14
(push 1 was a practice run and no data were collected) in Figures 1 - 12 in Appendix B.  The fluorescence
data are plotted as a function of the actual depth (offset from the tip depth) of the window utilized by each
channel.  An examination of the data shows that almost all push sites (except push #  14) contained some
degree of contaminant fluorescence.   This was by design; to  achieve our  goal  of demonstrating the
instrument's capability, we wanted to see as much contamination as possible.  A characterization of a site
depends as much on contamination non-detects as it does upon detects, for they define the boundaries of
the pollution plume.  In performing this field test of our instrument, we were interested in its response to
contamination and not accomplishing a site characterization. The push sites were chosen based upon the
amount  of contamination  that was expected  at a particular location, based on  monitoring well data
collected by Tom Best of the Air Force, our site coordinator.

     In  a general sense the molar absorptivities of classes of compounds can be summarized  for the
excitation wavelengths  used in our  system as follows: excitation wavelengths 257 nm, 266 nm, and 278
nm are known to excite BTEX compounds,  so any fluorescence in those channels might be attributable to
that type of species.  The channels with excitation wavelengths 289 nm, 299 nm,  315 nm are known to
efficiently excite fluorescence in naphthalene type compounds. Larger ring species will be most efficiently
excited by the excitation wavelengths 327 nm and 342 nm. These broad simplified categories  are not
exclusive. For instance, some three ring species such as anthracene absorb 266 nm light, but to a lesser
extent than they do at longer wavelengths.  These generalized categories provide some insight into the
trends seen from push to push.   A summary of the excitation wavelengths that correspond to the
maximum intensity channel in the most intense EEM of each push is given in Table 5. The peak intensity
and associated depth for each of the

Table 5. Summary of normalized summed peak fluorescence and depths.
  Push #  Maximum Summed Peak       Peak Depth (ft)          Excitation  X (nm)
                    Height
                 (relative units)
2
3
4
5
6
7
8
9
10
11
12
13
14
2.2E-08
2.4E-07
2.3E-07
4.1E-07
9.1E-07
1 .6E-06
1.8E-07
6.6E-07
1.0E-07
3.5E-06
1.8E-07
2.3E-06
No
18.1
13.4
12.0
7.1
12.1
16.0
13.0
13.0
13.0
9.6
9.6
12.6
No
289
327
266
289
299
266
266
266
266
289
299
299
No
                                              50

-------
pushes is given with the excitation channel responsible. The push with the maximum peak signal was #
11 and the lowest peak signal seen was in push 2.  The peaks arise largely from excitation wavelengths
that best excite naphthalene and BTEX type compounds. Of the 12 pushes that had contaminant signal, 5
out of 12 had their maximum peak due to 266 nm excitation, 6 out of 12 had their peak due to 289 nm or
299 nm excitation. Only one channel had a maximum signal in the 327 nm channel, perhaps indicating a
greater quantity of higher order PAHs at that depth.  The dominance  of naphthalene and BTEX type
compounds is consistent with our knowledge of the fluorescent compound composition of jet fuels such
JP4, the fuel spilled at this site.  In general, the peaks in the depth vs. fluorescence plots for the different
excitation channels correlate well with one another. The exact location of a peak along the  depth axis
may vary due to the specific composition distribution within the plume.

     Of interest to us in  terms of detection limits  and environmental  concern is the detection of
fluorescence due to compounds dissolved in the groundwater.  In terms of detection limits,  the water
solubility of BTEX compounds and PAHs is very low. Any signal arising from depths known to be below
the water  table indicates signal arising from small quantities of compounds in the zone.  In Figure 4,
Appendix B, the results for push 5 are shown for all of the excitation channels.  This push site was located
near monitoring well 10 (MW-10 in Figure 2).  The depths of the fuel product and the top of the water
table from the ground surface were determined to be 15 and 16.5 feet respectively. In Figure 4, Appendix
B, there is signal in several channels at  17  and even  18 feet.  While these signals are small, they  are
significant due to the  region in which they have arisen.   The water table depth  is known to vary
significantly over the course of a year and so there may very well be product that has been left behind on
the soil in the water table.  The signal seen is likely to be due to a smear zone below the water  level. The
depths of the fuel product and the water layer are not known for all of the push sites, but are known to
vary significantly from one push site to another.

3.  Representative EEMs
     A desirable goal of data analysis for this instrument is the complete automated EEM analysis of  the
field data. The rapid fitting of library (laboratory measured) EEMs to the in situ EEMs taken during field
work  (after correcting  the  latter  for soil effects)  would be  the ideal analysis scenario.  Even such a
sophisticated  analysis must be preceded by the type of analysis described in this section.  The summed
fluorescence vs. depth plots described in this section are not normalized, but the qualitative shapes and
peak positions are sufficient for comparison and discussion.23  In push site 6, seen in Figure 25, two depth
regions  of maximum fluorescence were seen with  peaks at 8.2 feet and 13.5 feet.  There are  significant
differences in the EEMs at the two depths.   At 8.2 feet, the fluorescence emission  is at much longer
                                               51

-------
wavelengths extending beyond the longest wavelengths measured by the detector.  This is in contrast to
the EEM taken at 13.5 feet where the fluorescence emission in the middle four channels begins at shorter
wavelengths and does not extend as far to the long wavelengths as that in the previous EEM.  This
provides evidence that the contaminants at 8.2 feet have a different chemical composition than those at
13.5 feet.  At this site, the old fuel  pipeline was thought to be buried just below our push site at
approximately 6-8 feet.  Our fluorescence data confirms this with signal beginning at 8 feet and extending
down to 10 feet. In addition, during the
            Signal (counts)
          -05  l.E+06  2.E+06  3.E+06
      Figure 25. Summed fluorescence vs. depth for push site 6 with peak EEMs.
                                                52

-------
            Signal (counts)
        l.E+05     5.E+06     l.E+07
      Figure 26. Summed fluorescence vs. depth for push site 11 with peak EEM.

push significant resistance was experienced and the cone tip pressure was elevated at around 8 feet. This
indicates that  we may have been pushing through the old  pipeline with residual fuel product in  the
vicinity.

     On the other side of the site, near the large above ground tank locations, we encountered the largest
signal at push site 11 seen in Figure 26.  For this peak contamination  region beginning  at 9 feet,  the
contaminant is fluorescing most strongly with 288.5 nm  excitation,  with a maximum intensity of
approximately 30,000 counts.   The high summed fluorescence indicates a large plume between 9 and 10
feet. Signal was seen at all depths below this plume, but at much lower light levels (500-1500 counts).
It is plausible that a foot of fuel product was resting on top of the groundwater and the signal seen below
that was due to groundwater contamination and/or  residual fuel product that  has adsorbed on the soil as
the groundwater level varied over time.
4.  Peak fluorescence EEMs

     An EEM from the depth of maximum fluorescence at each push location at the site was generated
and they are shown in Figure 27.  These EEMs illustrate the changes in the contamination at each push
location across the site at the  specified depth of peak summed fluorescence.  One EEM was generated at
the region of maximum fluorescence seen at each push location.  The EEMs vary significantly throughout
                                               53

-------
the push locations at the site, indicating a wide variety of chemical species present.  The peak EEMs from
each location can be grouped according to their respective EEM patterns. Push locations #3 - #9 all have
at least some contribution from a compound(s) with (A*, Am) maxima in the middle region, (290 nm - 300
nm, 350 nm). This correlates reasonably well with the EEMs of naphthalene and both jet fuels, indicating
the most likely contamination at these locations to be due to jet fuels.  At push locations #2, #5, and #11-
#13, the EEM patterns are comparable indicating similar fluorescent compound composition.  An overall
summary of the  site based upon EEM patterns allows the grouping of regions based upon contaminant
fluorescence,  and hence  type.   Variations  in soil  composition  and fuel  distribution will hamper  the
interpretation of our EEM data due to the 1.5" offset  of the  windows.  Even with EEMs that have been
reconstructed from data measured in successive exposures, small differences may  exist between the soil
regions probed by each window.
                                               54

-------
e
SB
    Excitation (nm)

                                                                Jf

                                                                '
                                                                                              310 a!o all 303
    Push 2,16 ft     Push 3,14 ft    Push 4,10 ft      Push 5,10 ft     Push 6,12 ft    Push 7,14 ft
    Excitation (nm)
e
SB
SB

                                                                                                   t
  Push 8,13ft      Push 9,12ft     Push 10,16 ft     Push 11,10 ft    Push 12,10 ft   Push 13,12 ft
 Figure 27. EEMs measured in situ at Hanscom AFB, Bedford, MA. The labels are the push #, EEM depth, and the plot axes are

 vertical: emission (nm), and horizontal: excitation (nm).

-------
5.   Depth corrected EEMs

     The  qualitative examination of in situ  measured EEMs can  yield information on the type of
contamination encountered.  The comparison of reference EEMs with field measured EEMs can indicate
regions of specific contamination such as JP-4, JP-5, and fuel oils.  Alternatively, regions can be grouped
according to contamination type, single rings (BTEX), double rings (naphthalenes), and larger species.

     The analysis of the peak EEMs, with maximum fluorescence signal, from each push location allows
the identification of the types and  sources of contamination present.  The EEMs in the following
discussion are all depth corrected, which involves constructing an EEM from  channels at the same depth.
The peak fluorescence EEMs from push locations #6, #7, and #11 are  shown in Figure 28, a and b, c and
d, e and f, respectively.

50-
1 -

6
w 25_




,~, 7 "
c^


•g '•









a


^IP- 	 ;::

Excitation A (nm)
| ™:::: -4(1

b








Excitation A (nm)
| ----- :^H
. i i i .


t::


W 425-




j:
c-;
c 350~

w "5
g 400-








c
£-f>



Excitation A (nm)
| =:::: -;f^H

d

-f;






Excitation A (nm)
i —-- >^m
i i i i . i


ii


•55 ""
W 25-




j:
i'335"


w "5
g 400-

«BO-






e


i

Excitation A (nm)
| =:::: -^^H

f






i

Excitation A (nm)
| ----- :iJjM
i i i i i i r























Figure 28. EEMs from the peak fluorescence depth regions of push locations a) #6: 6.5 ft + 0.2 ft, b) #6: 12.1 ft
+ 0.1 ft, c) #7: 14.1 ft + 0.3 ft, d) #7: 16.1 ft + 0.1 ft, e) #11: 9.7 ft + 0.1 ft, f) #11: 16.0 ft + 0.1 ft.
     The depths of the EEMs generated for the peak fluorescence regions at push location # 6 were 6.5 ft
± 0.2 ft and 12.1 ft± 0.1 ft.  There is considerable difference between the contamination seen at each
depth as evidence  by the EEMs, Figure 27,  a and b.  The contamination at 6.5 ft had an EEM with a
                                               56

-------
maximum at (315 nm, 415 nm), while the contamination at 12.1 ft had an EEM maximum at (300 nm,
350 nm).  The contamination at 12.1 ft has an EEM that is fairly similar to either of the jet fuels seen in
Figures 15,  a and b,  with EEM maxima at (289 - 299 nm,  340  nm).  At 6.5 ft, there may be some
fluorescence from species, such as naphthalene, found in jet fuels, but the majority of the signal is at
longer excitation and  emission  wavelengths indicating  the  presence of species which have  higher
conjugation. Consider the EEM of anthracene in Figure 14 c, with long wavelength excitation (299 nm -
342 nm) and  an EEM maximum at  (315 nm, 395 nm).  This  EEM pattern is  similar to  the long
wavelength excitation and emission found in the EEM measured at 6.5 ft. Furthermore, the characteristic
anthracene peaks can clearly be seen in the emission of the EEM from 6.5 ft at push location  #6.  The
smaller amount of short wavelength excitation (266 nm) similar to that in the EEM of anthracene can also
be seen in this EEM. In addition to anthracene, the  in situ EEM  indicates that there are other species
present. This is seen by considering that there is significant fluorescence intensity in the EEM from 450
nm -  500  nm which is not present  in the reference anthracene EEM.  This fluorescence EEM pattern
could be due to a compound such as pyrene (four fused rings), known to be excited most efficiently by our
315 nm channel and emit fluorescence from 370 nm - 530 nm.31  However, pyrene is only one possible
identity, as many compounds emit fluorescence in this region.

     The depths of the EEMs generated for the peak fluorescence regions at push location # 7 were 14.1 ft
± 0.3  ft and 16.1 ft ± 0.1 ft. There is a  similar difference between these EEMs, Figure 7, c and d, as there
was with those seen at push location #6.  The EEMs at shallow depths have a greater amount of long
wavelength  excitation and emission wavelength fluorescence  than do the  deeper layers.  This is  not
unexpected due to the proximity of the push locations, seen in  Figure 1. While the contamination depth
layers are different,  the characteristic patterns are comparable.  However, there is a dominance of jet fuel
type fluorescence in the  shallow layer at push location #7 and a smaller amount of long wavelength
excitation and emission wavelength fluorescence.

     The depths of  the EEMs generated for the peak fluorescence regions at push location # 11, on the
other  side of the  site, were 9.7 ft ± 0.1 ft and 16.0 ft ± 0.1 ft, shown in Figure 24, e and f.  The EEMs seen
at this push location are completely different from those seen at  either push location #6 or #7, indicating a
different type and source of contamination.  The EEMs at push location #11 are characterized by strong
excitation by the 289 nm channel and long wavelength fluorescence (325 nm - 500 nm). In addition, both
EEMs have some 266 nm excitation, but it is not seen in the first shallow layer due to the  overwhelming
signal in the 289 nm channel.  While there are different concentrations, indicated by the fluorescence
light levels, the contamination appears  to be the same at each depth.  This may indicate a single  source of
contamination in this site region and that the contamination has partitioned itself at two depths as  the
water table has varied from 9 ft to 18 ft below ground surface between 9/95 and 2/97. The variation was
                                               57

-------
determined via periodic  measurements made during this period with  an oil / water interface  probe
deployed in a well.

3)  Coast Guard Station, Elizabeth City, NC

     The site chosen by the EPA project leaders at RSKERL in Ada, OK for the second official test of our
instrument was the Elizabeth City Coast Guard Station in Elizabeth City, North Carolina, seen in Figure
29. A detailed map of the support center is shown in Figure 30.  The site was in the process of being
studied by conventional drilling and a combination of simple on-site analysis and standard analytical lab
analyses. This conventional analysis was being performed by another team from RSKERL.
                                            y.s.e.e.
                                        SUPPORT CEMTER,
                                                  crrr
Figure 29.   Map  showing  the  location of the  U.S.C.G.  support  center  in
Elizabeth City, North Carolina, and surrounding areas.

a)  Site Operations
     Upon arrival at the site on August 28, 1996, the field research team began working on the final
installation and testing of the system in the CPT vehicle and the assembly and testing of the generator
purchased to power our instruments in the field. After one and a half days, the system was operational
and the field measurements began.  There was some re-optimization of the system with fiber channel
reassignments during the latter portion of the first operation day.
                                               58

-------
    Fuel Farm
                    ttxy,
                                                               United Slates Coast  Guard
                                                                     Center,          CRy
Figure 30.  Detailed map of entire U.S.C.G. support  center in Elizabeth City, NC, with the area of CPT
investigation, the fuel farm shown in the upper left-hand corner.
The system had been set up for increased depth capability, by increasing the launch / detection cable by 5
meters.  This allowed the fiber optic connector junction box to be closer to the rod racks, enabling more of
the 20 meter cable to be threaded through the push  rods.  The efforts to  increase push depth were
successful and the maximum push depth attained in NC was 37 feet during the first push. This represents
a significant improvement over previous operation with 25 feet total push depth capability.  While  the
ability to push  to  such a  depth using  our  LIF  probe system  is  important,  the  apparent lack of
contamination at those  depths  indicates that the shorter fiber  optic cable  might have been  more
appropriate.  With shorter fibers, the light losses due to fiber optic attenuation are  less and therefore  the
system's detection limits  would be  lower.  The lengths of fiber used in this work were shorter than those
that can be used with either ROST or SCAPS LIF-CPT systems.  However, using the longer wavelengths
such as those used by the systems mentioned above would enable our system to probe the same depths if
desired.
                                                59

-------
     Several days into the field work, the laser head temperature sensor failed.  This was not a fatal
failure and turned out to be only a relatively minor setback (1A day), solved by replacing the temperature
sensor with one shipped to us by the laser manufacturer.  The field work then continued successfully:
Large contaminant signals were seen and the amount of signal observed was in agreement with available
information on where the contamination plume should extend.

     During the second week of the field test, we had a laser cooling water line rupture.  This failure, too,
was  easily remedied by purchasing and installing tubing  obtained locally.  This setback amounted to
another !/2 day of instrument down time.  Field data collection continued normally for the remainder of the
week until the system and  truck were packed and prepared  for the journey back to Massachusetts for our
third field exercise of the summer, again at Hanscom AFB.
b) Field Results

     The objective of our work at the U.S.C.G. Support Center in Elizabeth City, North Carolina, was to
demonstrate the instrument and characterize the extent and type of contamination at the  fuel farm.
Towards those goals, much was  accomplished with 27 pushes in less than two weeks.  Of the 27 push
locations, many were detects, but several were non-detects indicating the boundaries of the plume. The
LIF push locations, core sampled locations, and site features are shown in Figure 31.

     The stability of the  system during  the  field measurements was fairly good,  but  with room for
improvement.   The average deviation of the before  and after each push calibrant measurements was
usually better than 35 %.  However, there were several cases where one  channel underwent an order of
magnitude change. These drastic changes are most likely due to severe fiber damage or misalignment and
were corrected before the next measurement by re-polishing or re-aligning the fiber end.

     Normalized  summed fluorescence versus depth plots were generated using the excitation intensity
average for  each  push  location.   These plots were compared with plots of contaminant concentration
obtained from conventional laboratory analyses performed on  core  samples taken near some  of our LIF
push sites.   The  summed fluorescence data was used to generate  a 3-dimensional site map  for plume
visualization.
                                               60

-------
                            •14
          15 <
                                 >27
'16
                    >26
 •25

-------
1.  Calibration data

     The quantitative analysis of the field data will require a thorough knowledge of the variations of the
excitation energy delivered to the sample during the in situ measurements.  However, the exact incident
energy delivered to the in situ soil and contamination, for each measurement, is unknown.  The available
alternative is to measure  and record the  excitation energy before and after  the push  and from  these
measurements to infer the amount of excitation light  being delivered to the sample.  The calibration
process
and the treatment of the data are discussed and described  in section 5.B.2.b.l.  The excitation data in
Figure 32 show that there were significant changes in many of the channels throughout the field test.
Channels 258 nm and 266 nm both had fairly constant excitation energy levels, whereas channels 278 nm
and 288 nm had low output for the first 16  pushes, after which the energy level increased.  The remaining
channels showed more variation from push to push throughout the field test.  Several channels have large
error bars associated with the excitation energy average which can be due to  several factors previously
described.  The error in the excitation energy is propagated in the calculation of the  normalized summed
fluorescence data.  While a channel may have an error that  is an order of magnitude or larger, the signal
seen on that channel can usually still be  interpreted qualitatively to  determine the depth of contamination
but not the relative amount.
                                                62

-------
                             Excitation Energy vs. Push #
                                                     288.40 nm
                   - n ft
                            Calibrant Signal Sum vs. Push #      278 45 nitl
                          (quinine sulfate fluorescence sum only)        "
                             .i.a.a.
>
E? -S
iS f
I t
4 OE+10 -
3 5E+10
3 OE+10
2 5E+10
2 OE+10
1 5E+10
1 OE+10
50E+09
0 OE+00
Excitation Energy vs. Push # 266.03 IlHl
1 H R n R fl fl R ft n n n n 1 1 S 1 1 1
III!
3


Calibrant Signal Sum vs. Push # 2S7 4S HIT1
(quinine sulfate fluorescence sum only)
E
1 OE+05
8 OE+04
6 OE+04
4 OE+04
2 OE+04
_MJM
3 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26
Push Number

                                                                                                    Excitation Energy vs. Push #
                                       341.56 nm
                           .n.n.
                                                                                             6  7 8 9 10 11  12
                                                                                                               16 17 18  19 20 21 22 23 24 25 26
                                                                                                   Excitation Energy vs. Push #
                                       327.69 nm
                                                                                  a t
                                                                                     5 E+1
                                                                                     5 E+1
                                                                                     3 E+1
                                                                                     3 E+1
1 E+1
5 E+11
OE+00
illii
ft fl II ft. _ ft
                                                                                                    Excitation Energy vs. Push #
                                                                                                                            314.89 nm
                                                                                                           --§
                                                                                         3 5  6  7  8  9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26
                                                                                                    Excitation Energy vs. Push #       299 1 0
                                                                                  ? ?  10E*
                                                                                  I I  8 OE*
                                                                                             6 7  8  9  10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26
                                                                                                        Push Number
Figure 32. Excitation energy as a function of push number during the field test at the U.S.C.G. Support Center at Elizabeth City, NC.
Excitation channels 257.45 nm and 278.45 nm have no calibration data and are therefore expressed as calibrant fluorescence sums only.

-------
2. Depth vs. summed fluorescence

     The summed fluorescence data for each push location at the U.S.C.G. Support Station in Elizabeth
City, NC are given in Appendix B, Figures 1 - 27. In Figure 1, the data shown were taken "on the fly":
90 eight-channel LIF  measurements  in  15  minutes to  a depth  of  37 feet.   This push illustrates the
capability of the system to acquire  data rapidly during real time ASTM (2 cm/s) push rates. Since an
automatic data acquisition system was not yet in  place,  the depth data could not be used to trigger the
collection of each file.
Table 6. Summed fluorescence approximate peak depths and  contamination
depth ranges
Push*
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
na
nd
Peak depth average (ft)
na
8.03
9.00
8.11
9.06
8.08
9.05
8.05
5.05
4.04
4.02
8.05
nd
7.05
8.02
6.02
8.05
9.05
8.05
8.05
5.11
8.11
nd
6.05
nd
nd
nd
Not applicable
Non-detect
o Contamination range (ft)
-
0.08
0.01
0.10
0.07
0.09
0.07
0.06
0.06
0.16
0.07
0.06
-
0.06
0.07
0.07
0.06
0.06
0.06
0.27
0.10
0.10
-
0.06
-
-
-


-
6-12
4-13
6-10
3-11
5-9
3-9
7-13
5-7
3-6
2-5
6-11
-
6-7
4-9
4-9
4-10
3-11
5-9
6-12
4-6
5-9
-
5-8
-
-
-


                                              64

-------
Instead,  the plan was to obtain the time  each file  was  written and correlate  it with  time-depth
measurement  recorded by the  CPT computer, to obtain depth, but  this strategy was  unsuccessful.
Therefore, the depths were linearly approximated from 0 to 37 ft for the 90 data files. This push, without
accurate depth data, is not included in the other analyses.

     Measurements were made from window depths of approximately 1.5 ft to 16 - 20 ft, depending upon
the contamination plume depth.  In Table 6, a summary of the  average depth to peak fluorescence
intensity for each push location and the contamination zone depth (average of all channels) is given.  The
contamination had an average peak depth of 7 ft and the contamination depth range was from 3 ft to 13 ft.
The plots, in general, are self-consistent in that peaks in the different channels often occur at the same
depths.  This is expected if the contaminant mixture contains fluorescent species that  absorb at  all
excitation wavelengths, and if the mixture composition does not change significantly with depth.  While a
jet fuel fluoresces most strongly at central wavelengths, all of our channels respond to high concentrations
of fuels such as found in a saturated product zone.
3.  Summed fluorescence site map
     The normalized summed fluorescence data given in Appendix B have been plotted using the surface
x, y, z coordinates of the push locations at the site.  The surface elevation, z, was not measured, but was
nearly constant for all pushes, and was set equal to the average elevation, 2.4 m ± 0.3  m, of several other
points on the site to generate the 3-D site map in Figure 33.
                                                65

-------
     Northing (M)
                                                                                          Easting (M)
Figure 33.  Normalized summed fluorescence data, produced by the 288.4  nm excitation channel, plotted
with site coordinates to produce a 3-dimensional map of the fuel farm at U.S.C.G. Support Station at
Elizabeth Citv. North Carolina.

-------
The  3-D site map in Figure 33 was generated using Site View software.  This software uses all of the
summed fluorescence data points and creates a 3-D grid within which data points are linearly interpolated
to fill the 3-D grid.  A weighted sum interpolation algorithm is used to generate the data points that make
up the 3D site model.   In Figure  33, the parameters of the  weighted interpolation had been set to
emphasize the horizontal direction and not the vertical plume profile.  This allows an aerial-like view of
the site, push locations and contamination regions.  The summed fluorescence data suggest that there are
two main regions of contamination: the stronger in the middle of the sampled region of the site and a
slightly weaker plume center just northeast of the first.  The data in this summed fluorescence site map
were taken from the 288.4 nm excitation channel.  This channel was chosen because it produces the
maximum fluorescence response in JP-5 jet fuel (refer to section B.2.b).  If the jet fuel standard (Figure
14.b) is compared with the in situ EEMs from the U.S.C.G. Support Station in Elizabeth City, NC, all
except  one peak EEM contain at least some fluorescence pattern that is similar to JP-5. The axes on the
contamination intensity scale (color bar, Figure 33, inset) are from 10~9 to 3 xlO"6 fluorescence counts /
photons excitation.  In terms of JP-5 concentration, these correspond to 0.02 g / L (a non-detect) to 64 g /
L of JP-5 in cyclohexane on a medium grade sand.
4.   Summed fluorescence comparison with laboratory data

     Core samples were taken with a separate drilling rig after our work was completed,  at surface
locations near where CPT-LIF measurements were made.  The crew that collected these samples did not
duplicated  the  coordinates of our  push locations  as  closely as would be  necessary for  a detailed
comparison. In most cases, the  core sample locations were within a few meters of the LIF-CPT push
location.  An x-y plot of the push  locations, with site features,  is shown in  Figure  31.  The easting,
northing coordinates of the core sample sites and the LIF-CPT sites are given and compared in Table 7.
The coordinates for LIF-CPT and core sampling locations for push 2, 4 and 5 are separated to an  extent
that warrants concern about correlation.  For the remaining locations, the data should be similar, but
slight shifts in peak locations might occur due to variations in contamination plume depth and profile.

     The  laboratory analysis on core samples was  performed  by ManTech  Environmental Research
Services Corporation in conjunction  with the National Risk Management Research Laboratory (NRMRL),
Ada, OK.  The method used is the EPA Ada Laboratory's in-house standard method RSKERC SOP 72:
Quantitative Analysis of Aviation Gasoline  and JP-4  Jet Fuel in Coarse and Medium Textured Soils by
Gas Chromatography.  The analysis for TPH was done using an HP 5880 GC  (gas chromatograph)  and
the data expressed as equivalents of JP-4 jet fuel by using a standard fuel calibration curve. The analysis
                                               67

-------
of core samples for other specific compounds was performed using an HP 5 871A GC/MSD (Mass Spectral
Detection) system. The GC/MSD data were quantified using fluorobenzene as an internal standard.

     The laboratory data from core samples taken at the Coast Guard Center and comparable LIF data are
shown in Appendix C.  Figures  1-8 provide a comparison of laboratory core sample analysis data for
naphthalenes (naphthalene, 1-methylnaphthalene, and 2-methylnaphthalene) and total BTEX with in situ
LIF data.  Figures 9 - 16  are a comparison of TPH (total petroleum hydrocarbons) with the in situ LIF
data.  TPH data expressed as equivalents of JP-4 should be comparable to all channels in the LIF probe,
since each channel is sensitive  to jet fuel.  The comparison  of specific compounds is more complex
because it requires some interpretation of the  EEM data. For our system, it is known that naphthalene is
only excited by the 266 nm, 278 nm, 289 nm, and 299 nm excitation channels,  with the latter two having
the most intense response.  The same should be approximately true of the naphthalene derivatives.

Table 7.  Summary of LIF-CPT push locations versus core sample locations for lab analysis.

Push name       Easting (m)   Northing (m)  Qualitative Agreement Between Lab
                                                       and LIF Data
Push 2
cpt-2
Difference
Push 3
cpt-2
Difference
Push 4
cpt-2
Difference
Push 5
cpt-1
Difference
Push 6
cpt-3
Difference
Push 7
cpt-7
Difference
Push 9
cpt-4
Difference
Push 10
cpt-6
Difference
862826
862827
0.7
862829
862827
2.3
862837
862827
10.3
862843
862845
2.0
862869
862867
1.6
862800
862800
0.4
862868
862868
0.1
862909
862909
0.2
283120
283111
8.7
283113
283111
1.7
283111
283111
0.3
283125
283128
3.4
283157
283158
1.2
283168
283168
0.3
283072
283072
0.3
283069
283069
0.2
Poor

Fair

Fair

Good

Good

Fair

Good

Poor-Fair

                                              68

-------
are only excited by our lowest two excitation channels: 257 nm and 266 nm. Of these two, 266 nm is the
stronger due to a  higher energy before launch and less fiber optic attenuation before delivery to the
sample. To assist in the correlation of LIF-CPT measured signals with the core sampled lab data, straight
lines are drawn in the figures at the depth corresponding to the peaks in the lab data across the in situ
measured data.

     In general, there are summed fluorescence peaks at similar depths to the peaks from the lab core
sample analyses. It is difficult to closely correlate these measurements due to their inherent differences.
First, our fluorescence data originate from a small sample area (< 1 mm) dictated by the excitation fiber
distal cone cross-section on the outside surface of the sapphire window.  This is in contrast to lab analyses
where a 4 inch core section is homogenized and a sample extracted with methylene chloride. Second, our
data are plotted as summed fluorescence at each  excitation channel and the lab data is in  concentration
units for groups of specific  compounds (naphthalenes, total benzenes) and  for TPH (total petroleum
hydrocarbons).  For the TPH measurement, the only fluorescent compounds that the analysis detects are
BTEX and naphthalenes.

Compound comparison

     For LIF-CPT push # 3, the closest core sampling was 32CPT-2, seen in Figure 2, Appendix C. Both
the BTEX and naphthalenes contamination ranges correlate with the LIF data.  The lab data indicate both
contaminants from 5 ft - 9 ft, and the LIF data on the 266 nm channel (sensitive to BTEX)  begins at 5 ft
and ends at  13 ft.  Similarly, the 289 nm and 299 nm channels have signal from 6 ft - 12 ft indicating
good overall contamination region  overlap.  The total  BTEX peak  at 6.5  ft correlates well with a peak
region in the 266 nm summed fluorescence plot.   The naphthalenes peak  is approximately 1A foot away
from the peak region on the channels most  sensitive to naphthalenes, 289 nm and 299 nm.  However, the
LIF data suggest contamination from 10 ft  - 14 ft, whereas the lab data do not show contamination from
10 - 12 ft, the end of their data range.  For both pushes #2 and #4, Figures 1 and 3 in  Appendix C, the
data correlate well for the overall contamination ranges, but the LIF push  and core sample locations are
farther apart, making the correlation less meaningful.

     For LIF-CPT push # 5, the closest core sampling was 32CPT-1, seen  in Figure 4,  Appendix C.
There is good correlation at this location, with a contamination region from 5 ft - 11 ft.  Both the lab data
and the LIF plots indicate a  shallow region of contamination at approximately 2.5 ft - 3 ft. At LIF-CPT
push # 6, the core sampling was 32CPT-3, seen in Figure 5, Appendix C. Both the overall contamination
region and the two main peaks in the lab data are well correlated with the LIF data.  The lab data show a
                                               69

-------
region of contamination from 5 ft - 8 ft and the LIF data range is from 6 ft - 9 ft.  The two main peaks in
the naphthalenes plot overlap closely with those in the 289 nm and 299 nm LIF channels.

     For LIF-CPT push # 7, seen in Figure 6,  Appendix C, the lab data indicate shallow contamination
from 2 ft- 5 ft, while the LIF data show contamination from 3 ft - 7 ft. LIF-CPT push # 9, seen in Figure
7, correlates well with the lab data, with contamination beginning at 4 ft and ending at 7 ft. Finally, LIF-
CPT push # 10, seen in Figure 8, shows  some shallow contamination from 3 ft - 6 ft and 266 nm signal
from  the  surface down to  6 ft.  This is  comparable to the lab data which indicates  surface BTEX
contamination down to 3.5  ft, and naphthalene contamination from 2 ft - 3 ft.  The fluorescence signal
levels during this push were low, consistent with the low concentrations seen at this location.

TPH comparison

     For LIF-CPT push # 3, the closest core sampling was 32CPT-2, seen in Figure  10, Appendix C.  The
TPH contamination range correlates with the LIF data in the range of 5 ft - 11 ft. However, the LIF  data
indicate contamination from 5 ft to 14 ft, 3 ft beyond that reported by the laboratory. The peak depth in
the lab data lies between two peak regions in the LIF data. For LIF-CPT push #  5, seen in Figure 12,
Appendix C, there was very good correlation. The lab data shows contamination from 2 ft to 11 ft, while
the LIF data indicates contamination from 3 ft to  12 ft. Several peaks in the lab data correlate reasonably
well with  the LIF data.  The remaining data are given in Figures 13 -  16, Appendix C. The results are
similar to  that seen for the specific compound analysis comparisons.

     The  correlation  of the in situ  measured LIF data with conventionally  sampled  and  analyzed
laboratory data has  been generally good, with several LIF pushes in close agreement with the lab data.
The fundamental differences in sampling, data interpretation  / presentation, and actual  push vs.  core
sample locations prevent a detailed, quantitative comparison.
5.   Peak fluorescence EEMs

     For each push location, an EEM was generated at the peak summed fluorescence depth.  Each probe
window is located at a different depth relative to the cone tip during the push, and thus "sees" a different
sample. To correct for this, data from several EEMs at different depths were taken and each channel
assigned its depth.  A new EEM was created by collecting only channels close to the desired depth.  These
EEMs are effectively depth corrected to within 2 % depth variation.  In Table 5, the peak fluorescence
                                               70

-------
depths are the actual EEM depths with standard deviations. In Figure 34, EEMs are shown from the 21
detects encountered at the U.S.C.G. Support Station in Elizabeth City, NC.

     The EEMs in Figure 34 document the changes in the fluorescent contamination profile at the depths
of maximum fluorescence depths at the locations sampled.  The EEMs contain fluorescence patterns that
are  indicative of jet fuel type contamination with two notable exceptions: push # 10 and #11.  Many
EEMs that contain jet fuel contamination fluorescence signals also contain other patterns of fluorescence
indicating other sources of fluorescence.  The standard EEMs of both JP-4 and JP-5, Figure 14 a and b,
have fluorescence excitation maxima at 299  nm and 288 nm respectively and emission maxima occur at
325 nm - 350 nm. This fluorescence EEM pattern agrees  closely with the observed EEMs in Figure 33.
A common additional feature of the in situ measured EEMs, compared with the reference jet fuel EEMs,
is a tail of long wavelength emission.  This additional fluorescence emission in the 288 nm or 299 nm
excitation channels extends from 350 nm to 450 nm or 500+ nm. The two notable exceptions are push #
10 and #11, where the emission occurs at long excitation and emission wavelengths.  This EEM pattern
of emission is indicative of highly conjugated aromatic species such as anthracene (3 - ring structure) and
higher.  The EEM of anthracene is shown  in Figure 15,  c, with emission from 375 nm - 425 nm at
excitation wavelengths of 314 nm - 342 nm.   This provides evidence that the contamination, at  the
indicated  depths  and  push  locations,  is likely to  contain conjugated aromatic molecules  such as
anthracene.  These EEMs are valuable tools in the characterization of the contamination profile of the site
and indicate regions that can be grouped together based upon contamination type.
                                              71

-------
Push 2, 8 ft
 Push 8, 8 ft
Push 15, 8 ft
Push 21,5ft
Push 3, 9 ft
 Push 9, 5 ft
Push 16, 6 ft
Push 22, 8 ft
Push 4, 8 ft

Push 10, 4 ft
Push 17, 8 ft

                                     -«2
                                           Excitation A, (nm)
Push 24, 6 ft
Push 5, 9 ft
Push 6, 8 ft
Push 7, 9 ft
Push 11, 4ft
Push 12, 8 ft
Push 14, 7 ft
Push 18, 9 ft
Push 19, 8 ft
Push 20, 8 ft
Figure 34. Depth corrected EEMs for each push location at U.S.C.G. Support Center, Elizabeth City, NC. The plot axes are vertical:
emission wavelength (nm), horizontal: excitation wavelength (nm), and the labels are push # and depth.

-------
5)  Hanscom Air Force Base II

     The third field test field test of the summer of 1996 was a second visit to the site described in the
Hanscom Air Force Base I section above.  A return visit of the instrument and crew was arranged to
enable us to further test our instrument and hopefully accomplish a site characterization in the process.
The site is  shown in Figure 35 with site features and both the first (Hanscom I) and second (Hanscom II)
push locations.
                                           10'      9 O
                                                             8O
                                           3 •     4 '       5 '
                                        MW-9 O           MW-10 O
     „   Former railway line for
MW-11O   fuel delivery
                                                   11O
                                                               12O
                      Former
                      underground
                      fuel
                      pipeline
                                                  MW-17O  MW-16O
                                                                                       18O
                                          Fuel recovery system
                                                   14"
                                                                                     200
                                      FormerASTs
Figure 35.   Map of CPT-LIF push locations and site features at Hanscom AFB.   The dots are the push
locations from the first field test, and the diamonds  are the push locations from the second field test.  Push
location # 15 was crossed out because a shallow concrete slab prohibited further progress.

a)  Site Operations
     Several organizational changes were made to the layout in the truck once it had returned to Hanscom
AFB. There were no major modifications or repairs attempted due to lack of time.  In addition, repairs to
an essentially operational system would be problematic in that the results taken with the modified system
would not be directly comparable with those taken during the previous work.  The organizational changes
                                                 73

-------
included the removal of the laser shipping case from the truck, securing the laser directly to the rear cab
floor, and repositioning of the computers for easier access.

     The CCD camera had to be replaced due to a vacuum leak in the chip chamber and a Peltier cooler
malfunction.   Luckily,  an  essentially identical detector  was available in our laboratory and this was
quickly swapped with  the damaged unit after a few modifications.  The re-installed  system was tested
prior to the arrival of the EPA crew and appeared to function normally. However, just prior to the first
push, the laser overheated and shut down - the cooling group fans were not functioning.  This problem
was not trivial and required l!/2 days to repair due to the time needed to ship and replace the part (Fan PC
board).  This action corrected the problem and the field work was begun during the second day.

     Having previously done field work at the  site, we embarked on this  second visit with hopes of
accomplishing  a complete  characterization,  with preconceived  notions  of locations  and  levels of
contamination. We first tried to duplicate our earlier results at several push sites where we had seen large
amounts of signal previously. These measurements were initially perplexing because the results were very
different from what we had seen before.  No one, including the site manager, Tom Best, expected to find
such drastic changes. The absence of substantial fluorescence signal was a source of much concern for all
involved. Nevertheless, the instrument response from all calibrating solutions (Quinine sulfate, JP5, and
fuel  from  a  nearby well mixed with  sand) was as expected.   The  conclusion was that  the site
contamination had changed since we had sampled there three months earlier.  We proceeded with the site
characterization.  On the fourth day of sampling a highly contaminated subsurface zone was discovered.
Its location was consistent  with  the  site manager's expectation: near and in  between  two  pumping
recovery wells.  In the four days of field sampling, 21 surface locations were probed to depths of 25 feet.

b)  Field Results

1.  Calibration data

     The performance of the LIF instrument can be directly monitored by the fluctuations in the response
of the calibrant, quinine sulfate, before and after every  push.  For the first 6 pushes, 8 channels were in
use, until the 266 nm excitation fiber was damaged at the  probe end and it was swapped with the 278 nm
excitation fiber.  From push 7 on,  there were only 7  channels:  the  original  eight minus the 278 nm
channel. This is seen in Figure 36, with excitation energy as a function of push number during the field
test.  The data had larger errors during the first half of the field test, until push  11 or 12.  This may be
attributable to cold weather (outdoors 5-10  degrees Celsius) hampering the performance of the laser.
                                                74

-------
The solution was to allow significantly more warm-up time for the laser (1 hour +), but this was not done
until the problem
                                               75

-------
                           Excitation Energy vs. Push #
                                                   288.40 nm
   20E+12

I? 15E*12

|f 10E.12


II —"<
   0 OE+00
                                         ill




^
Ej
s
E
w
c
55

Calibrant Signal Sum vs. Push # 278.45 Iim
(quinine sulfate fluorescence sum only)
2E+05
2 E+05
1 E+05
1 E+05
1 E+05
8E+04
6E+04
4 E+04
2 E+04






f| 1
ll X R FJ T
11 PI II II 1 A









1 2 3 4 5 6 7 8 91011121314161718192021
                           Excitation Energy vs. Push #
                                                   266.03 nm
          I!
                 12345
                          Calibrant Signal Sum vs. Push #      257 45
                        (quinine sulfate fluorescence sum only)       "
                            7 8 9  10 11 12 13 14 16 17 18 19 20 21
                            Push Number
                                                                                                  Excitation Energy vs. Push #
                                                                                                                341.56 nm
I   §^
                                                                                                  Excitation Energy vs. Push #
                                                                                                                              6Q HIT1
                                                                                                  Excitation Energy vs. Push #
                                                                                                                          314.89 nm
                                                                                                              J..I.J.H.B.H.
                                                                                                  Excitation Energy vs. Push #
                                                                                                                299.10 nm
                                                                                                Jli
                                                                                2  3 4 5  6  7  8 9 10 11 12 13 14 16 17 18 19 20 21
                                                                                         Push Number
Figure 36.  Excitation energy as a function of push number during the field test at Hanscom AFB in Bedford, MA. Excitation channels
257.45 nm and 278.45 nm have no calibration data and are therefore expressed as calibrant fluorescence sums only.

-------
became apparent during push post processing. For excitation wavelengths, 257 nm, 288 nm, and 299 nm,
the excitation energy did not show significant long-term degradation over the course of the field test. This
is  in contrast to excitation wavelengths, 315 nm, and 342 nm, where the excitation energy tapered off
significantly. For 266 nm and 327 nm excitation channels, the excitation energy increased as the field
test progressed.  There are many possible reason for the changes in the excitation energy throughout the
field test.   These include SRS (stimulated Raman scattering) variations due  to  Raman  shifter  gas
decomposition or window damage.  Damage to excitation  beam launch  fibers and hardware are other
possibilities, such as launch tip damage, and fiber solarization.  A very important possibility is damage to
either excitation or emission fiber ends in the probe.  In either case, the signal returning to the detector
will be reduced either through loss of excitation light by the excitation fiber or reduction of fluorescence
collection by the emission fiber.  The large order of magnitude changes associated with some excitation
beams  are most likely due to catastrophic damage to the fiber end or changes in the  launch hardware.
Such large  excitation energy  variations  are accounted for in the error propagation of the summed
fluorescence plots.  Those excitation beams in each push with unacceptably large error bars can be
discarded from any quantitative analysis.  However, for qualitative purposes, the in situ fluorescence seen
by the probe  should not be  discarded.    These data indicates the  presence  and location of actual
contamination, and perhaps even identity, but not quantity.
2.  Depth vs. summed fluorescence

     The normalized, summed fluorescence data at each excitation wavelength for each push location are
shown in  Appendix D, Figures 1 - 20.  The data are summarized in Table 8, with peak fluorescence
depths and ranges of contamination.  Surface locations were probed to depths ranging from 13 ft  - 15 ft
below the  ground surface.  Several pushes were hampered by a hard sub-surface formation, threatening to
damage the probe and optics.  In particular, push 15 had barely begun when an underground concrete slab
was encountered, ending the push prematurely at 4 ft. The majority of push locations were ended at 18 ft
- 20 ft due to a hard glacial till that was impenetrable.

     The average depth of contamination, indicated by fluorescence, was 12 ft, with an overall range of 4
ft -  17 ft.  There were 11 non-detects (push locations in which there was  no significant contaminant
fluorescence) out of 21 total push locations. At the locations where signal was encountered, there was
partial correlation between the response in the channels, suggesting contamination that has a wide variety
of species  that absorb and emit on all channels used in the probe.  However, there are some depth regions,
in certain  pushes, where one or more channels do not agree, indicating a contamination region that does
not have as broad a range of contaminant species.
                                               77

-------
Table 8. Summed fluorescence approximate peak depths and  contamination
depth ranges during Hanscom II.
Push*
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
Peak depth average
nd
10.0
nd
nd
5.0
nd
nd
nd
10.1
nd
12.1
13.1
nd
11.0
nd
16.0
14.8
14.6
nd
11.1
nd
0
0.16
0.07
-
0.27
-
0.27
0.09
-
0.16
-
0.08
0.19
0.10
-
0.09
-
Contamination range (ft)
5-11, 16-17
5 - 8, 14 - 16
-
6-13
-
10-14
10-15
-
10-12
-
14-16
5-6.5, 14-15
14.5-15
-
11-12
-
    nd            non-detect
3.   Peak fluorescence EEMs

     An EEM was generated, for the peak fluorescence region, at each push location where fluorescent
species were detected. The EEMs were generated by labeling each detection channel with its actual depth
and then collecting individual channels from the same depth from separate acquisition files. The depth
corrected EEMs from those push locations with contamination are shown in Figure 37. The depths of the
corrected EEMs are given in Table 9 and are generally within 3 % of the given depth. The first two peak
EEMs, push # 2 and # 5, had fluorescence EEM maxima at (^^n) of (289 nm, 440 nm) and (342 nm,
445 nm) respectively. These two EEMs have very different peak regions.  However, the peak region  of
                                              78

-------
     340    320
                   2«0    250
        Push 2,10 ft
       Push 12,13 ft
                 Jp
       Push 18, 15 ft
a
o
                             S 4,
       Push 5, 5 ft
                                              1
                                       320 310 300  290 280 27'
        Push 14, 11 ft
                                      Excitation A, (nm
       Push 20, lift
                                                                          1
Push 9,10 ft
                                                                   320 310 300  290 280 27i
Push 16,16 ft
Push 11,12 ft
 Push 17, 15 ft
Figure 37. Depth corrected EEMs measured at Hanscom AFB, Bedford, MA. The labels are the push #, EEM depth, and the plot

axes are vertical: emission (nm) and horizontal: excitation (nm).

-------
fluorescence in push # 2 was also present in push #5, but to a lesser extent.  Both EEMs had fluorescence
emission in the 289 nm channel ranging from 325 nm to 500 + nm.  For the remaining 8 peak EEMs, the
maximum EEM fluorescence was seen on the 289 nm channel. The range of fluorescence emission was
from 325 nm to 500 + nm, with a peak ranging from 335 nm to 450 nm. When the peak of the EEM is at
(289 nm, 335 nm), such as in push # 11, 12, 14, the EEMs are similar to the reference naphthalene EEM,
Figure 15, b.  While the peak region is similar, the in situ EEMs contain a significant amount of signal
due to other species (emission from 350 nm to 500 + nm).
4.   Monitoring well sample

     A pure fuel product sample was taken from monitoring well 9 (MW-9), seen in Figures 23 and 35.
The EEM of that sample mixed with EC USCG sand is shown in Figure 37.  This sample is characterized
by low to middle wavelength excitation (266 nm - 315  nm), with middle to long wavelength emission
(325 nm - 500 nm).
       s  ^
       I  4oo^
      'S  425^
                340   330  32n  "In   J""  2"   _<-"  270  260
                          Excitation ^ (nm)	
                     I	  	v—a^^H
Figure 38.  EEM of monitoring well # 9 sample from Hanscom AFB, measured on
a medium grade sand.
                                             80

-------
In addition, there is some long wavelength excitation in the 341.2 nm channel with long wavelength
emission (380 nm - 500 nm). A comparison of this EEM with the reference naphthalene EEM, Figure 15,
b, and JP-4 and JP-5 EEMs, Figures  14, a and b, indicates some similarities.  One would expect the
reference samples (naphthalene, JP-4 and JP-5) to compare in a similar fashion due to the presence of
naphthalene fluorescence in the jet fuels. The naphthalene EEM has an excitation peak at 288.4 nm and
an emission peak at 335 nm compared with an excitation peak at 288.4 nm and an emission peak at 340
nm for the EEM of the well sample.  Also, the shape of naphthalene peak is consistent with  the peak
pattern seen in the well sample EEM. It should be noted that the well sample has significant fluorescence
extending from the  naphthalene fluorescence region to longer excitation and emission  indicating other
types of contamination.  The JP-4 EEM has significant signal arising from both the 288.4 nm and 299.1
nm excitation channels which also correlates with the well sample EEM.  The JP-5  EEM correlates with
the well sample in the naphthalene fluorescence region, but does not have the fluorescence in the 341.2
nm excitation channel that JP-4 EEM does. By comparison with the reference EEMs, it  is likely that the
well sample contains naphthalene and JP-4 type contamination with additional fluorescence due to other
species, probably three ring or larger due to the middle wavelength excitation / long wavelength emission.
                                               81

-------
VI. SUMMARY
A. Overview of Goals Accomplished

     The goals that have been achieved include the development and field testing of our fiber optic LIF-
EEM CPT probe, software development for data analysis, hardware and software schemes used for data
acquisition and data processing.  In addition, LIF-CPT measurements have been carried out at a well
characterized site,  and short term changes (3 months) at a remediated site have been monitored.  The
detailed calibration of the instrument to include instrument functions and artifacts has been undertaken
and  accomplished  at the summed fluorescence level.  The data  and calibration  schemes are general
enough that calibration of the  full EEMs  can easily be  accomplished  with software to perform the
normalization pixel by pixel.  Several reference compounds and mixtures have been measured in a variety
of media,  allowing comparison  with many of the field - measured EEMs.  The  fiber optic probes have
been evaluated quantitatively and tested on various media of different particle sizes to  determine response
changes.
B. Detailed Summary of Goals Accomplished in This Work

     As far as the hardware is concerned, the development of our instrument is nearly complete.  The
laser, optical table, spectrograph, detector and other equipment have been successfully packaged for and
operated in the field in RSKERL's cone penetrometer vehicle, which, with its separate hydraulics and
instrumentation compartments is somewhat roomier than the typical SCAPS or commercial CPT vehicle.
The instrument fits comfortably in the space allotted, but would have  some difficulty squeezing into the
latter.  Running the instrument  off the truck's onboard  generator is inconvenient, since this must  be
turned off between pushes.  The problem was solved by the purchase of a commercial gasoline-powered
generator, which is mounted on wheels and  is  dragged alongside the truck during between-push
repositioning by CPT and/or LIF personnel.  A system should be devised to mount the generator on the
front or rear of the truck to obviate the difficulties and dangers of this arrangement.

     The laser/Raman shifter has demonstrated potential as a multi-wavelength light source, but it has not
achieved a desired level  of stability in  output power in  the various  channels.  There has been some
difficulty in keeping the laser  cool enough in warm  weather:  proper  functioning of the truck's  air
conditioner units would help considerably. We would hope to bring fluctuations in short-term intensity in
                                               82

-------
each channel down to 10-20% if possible, and develop the capability to provide instantaneous photon
normalization (during the push rather than before and after).

     The use of the system with optical fiber cables and probe hardware to deliver excitation light in the
field has been demonstrated during three separate field tests made during the latter part of 1996.  The
system performed well and was able to collect data for a significant number of pushes at all sites.  There
exists some difficulty with both excitation energy stability during some of the pushes and stability from
push to push.   The stability from push to push is accounted for by the calibration  procedure, provided
there  is signal to calibrate.   However,  stability  during  a push  remains uncorrected and  we rely on
measurements before and after the push to indicate if constant excitation power levels were attained. An
examination of the error bars of each push average indicate that many of pushes had excitation energies
that were within  acceptable limits.   The histogram of the number of measurement versus CV (%) for
individual excitation energy measurements during both the Hanscom II and the North  Carolina work is
shown in Figure 39.  This plot indicates that the majority of the before and after measurements, 71 % (229
/ 323), were within  +40 %  of one  another.   There are  many options for the improvement of system
stability and these are discussed in Chapter 6.

     The fiber cabling arrangement and probe design proved to be straightforward and functional.  The
multi-window arrangement, however successful in avoiding cross-talk from channel to channel,  makes
reconstruction of  depth-corrected EEMs difficult.  This point is addressed again in the following section.

     The  first field  test at Hill AFB was perhaps slightly premature from the point of view of our
readiness, but it accelerated our overall progress and gave us some much-needed field  experience.  It is
difficult to assess the instrument performance during that test because subsequent site characterization by
another LIF-CPT crew indicated it was relatively uncontaminated.

     In subsequent field work, the  hardware performed well as indicated by a  number  of different
measures: transportability; good integration into vehicle and CPT system; reasonably  short installation
and mobilization  times; ability to function and generate both  calibration  data and  soil characterization
measurements nearly  continuously during scheduled  field  tests;  reasonable productivity in terms  of
number of pushes per day; success in demonstrating the ability of all channels to generate useful response
levels from contaminated soils; qualitative demonstration (clear in visual presentations of the data) of the
utility of multidimensional fluorescence  measurements in comparing contaminants to each other or  to
reference chemical compounds and mixtures.
                                                83

-------
    100T
 S3
 4>
     so-
     so-
                  87
             53
                      56
     401
 E

I
     20-
                           33
                                19
                                     16
                                         12
                                              10
                                                   11
             0.0       20.0      40.0       60.0      80.0      100.0      120.0      140.0     160.0

                 10.0       30.0      50.0      70.0       90.0      110.0     130.0      150.0



                                    Excitation Energy CV (%)



       Figure 39. Number of measurements versus the excitation energy CV for individual excitation

       channels over all pushes for the North Carolina and the Hanscom II field work

-------
     The postponement of our subsequent field work until late in the third year of the project, and the
intense field work schedule from June to October of that year, resulted in a very limited opportunity for
between-test improvements  and development of field-data analysis schemes. Our efforts during the six-
month no-cost extension were  split between implementing the automated data collection routine and
trying to develop suitable data reduction schemes for the large amount of field data we had accumulated.
It is fair to say that the data analyses accomplished by the end of the project are rudimentary, compared to
what we feel the quality and quantity of the field EEMs justify.

     To be more specific, the analyses we have described herein fall into two categories: 1) presentation of
the fluorescence data  in each channel  in  ways  similar to those used in single-channel  LIF-CPT
instruments; and 2)  largely  qualitative and visual inspection of the field EEMs, including comparisons to
reference EEMs generated on the same instrument.   The first category was pursued at the request of our
project manager and others who  wished to have  a simple, real-time display  available  during data
acquisition.  Its accomplishment demonstrates our instrument as a multichannel version of previous LIF-
CPT instruments.  The second category was the shortest, simplest way for us to demonstrate that the
multiple excitation channels of our instrument, taken as a whole, represent more than merely the sum of
their parts, and open the door to three-dimensional pattern recognition and other EEM data analysis
techniques.  A quantitative or  semiquantitative treatment of the field EEMs  that makes full use of the
mathematical machinery for analysis of matrix-formatted data is clearly desirable, and constitutes a major
focus of our current and proposed future work.
The comparison of in situ LIF  data with "true" laboratory data can be problematic in that the truth of
laboratory analyses can be compromised.

C. Some Accomplishments of Related Concurrent Work

     During this project, we also had a grant from NHSRC  for laboratory work in support of the field
demonstration, and  a demonstration grant from AATDF (DOD-funded through Rice University)  for a
second-generation field instrument.    Finally, another  multi-investigator grant from NHSRC helped
finance the field work at Hanscom AFB.   Some results  from these other projects that impact on our
assessment of the technology are given in this section.

     Because the provision  for more than one excitation wavelength adds complexity and cost to various
aspects of the instrument, as well as increasing the size of the data sets and the computational time for
any data processing, we undertook a limited study to determine what the analytical benefits of additional
excitation wavelengths might be.  We attempted to analyze two mixtures, one  containing four BTEX
compounds,  benzene,  ethylbenzene, toluene and p-xylene, and one containing a benzene, naphthalene,
                                               85

-------
and  anthracene,  simple  representatives  of the  one-,  two- and  three-ring aromatic molecules.   For
simplicity, the mixtures were prepared in cyclohexane and the spectra were measured on a conventional
laboratory fluorimeter.  Full EEMs of the mixtures were obtained and rows and columns were eliminated
to produce EEMs of various sizes, corresponding to 14, 10,  7 and 5 excitation wavelengths.  In brief, the
results were  that, using least-squares analysis to determine  concentrations, we could not  accurately
discriminate  among the  BTEX compounds even with the maximum 14 excitation wavelengths;  by
contrast, even with five excitation wavelengths the different ring sizes could be discriminated reasonably
accurately.  From this result, we have revised our original suggestion that individual compounds might be
identified.  While this might still be true for some  analyte with a very distinctive spectrum, it is not
generally the case.   On the  other hand, the results  also make a strong case for being  able to identify
classes of similar chemicals.  The task before us is to develop some tools for sorting target analytes into
classes and doing some studies to determine the quantitative  accuracy of analyses based on these classes.

     In developing a  second version of our LIF-EEM  CPT instrument,  we attempted  a number of
improvements.  We decreased the footprint of the instrument to effect a comfortable fit into a commercial
CPT vehicle,  we increased the laser repetition rate from 20 Hz to 50 Hz, with a corresponding decrease in
exposure times for each EEM, and we developed a new probe design that puts all ten channels in the  same
approximately 1 cm diameter window.  This last change solves the problem of different beams sampling
different depths, although each still looks at a different spot of the soil.  The new probe apparently has
increased cross-talk among channels, although this problem may be because the imaging spectrograph on
the new system is not as aberration-free; a complete analysis is still in progress.

     In the area of data analysis, we have  developed software that  should permit summed fluorescence
results to be available in the field.  We have reviewed recent developments in rank annihilation analysis
and incorporated the more powerful ones into our data analysis schemes.  We have also begun addressing
the full correction of all instrument artifacts necessary to  compare field data to a  reference library of
standard EEMs.  This  includes developing  a theoretical model for fluorescence intensity appropriate for
our probe geometry and applicable beyond the optically dilute limit.  We are nearing completion of our
own reference library of EEMs of aromatic molecules in various solvents as well as on different soil types.
                                               86

-------
                                        REFERENCES
Note: A * denotes a publication resulting from this project.

1.   "Remote Analysis of Groundwater Contaminants Using Laser Fluorescence", W.A. Chudyk and J.E.
    Kenny, Proceedings of Fourth National Symposium  and Exposition on Aquifer Restoration and
    Ground-Water Monitoring, 235 (1984).

2.   "Remote Detection of Groundwater Contaminants Using Far-UV Laser Induced Fluorescence", W.A.
    Chudyk, M.M. Carrabba, and J.E. Kenny, Anal. Chem., 57, 1237 (1985).

3.    "Prototype Laser Fluorescence/Fiber Optics Groundwater Contaminant Detector",  W.A.  Chudyk,
    M.M. Carrabba, G.B. Jarvis, and J.E. Kenny, Proc. Specialty Conf. on Environ. Eng.,  EE Div., 1985,
    p. 98.

4.   "Monitoring of Ground-Water Contaminants Using Laser Fluorescence and Fiber  Optics", W.A.
    Chudyk, J.E. Kenny, G.B. Jarvis, and K. Pohlig, InTech 34, (5), 53 (1987).

5.   "Remote  Laser-Induced  Fluorescence Monitoring of  Groundwater  Contaminants: Prototype  Field
    Instrument", J.E. Kenny, G.B. Jarvis, W.  A.  Chudyk, and K.O.  Pohlig, Analytical Instrumentation
    16, 423 (1987).

6.    "Instrumentation  and Methodology for Multicomponent Analysis  Using In Situ  Laser-Induced
    Fluorescence", J.E. Kenny, G.B.  Jarvis  and H.  Xu, Proc., First  International Symposium  on  Field
    Screening Methods for Hazardous Waste Site Investigations, 133 (1988).

7.   "Laser Fluorescence EEM Instrument for In Situ Groundwater Screening",  T.A. Taylor, H. Xu, and
    J.E.  Kenny, Proc., Second International  Symposium on Field Screening Methods  for Hazardous
    Waste Site Investigations, US.EPA, Las Vegas, NV (1991).

8.   "Laser-Based Fluorescence EEM Instrument for In-Situ Groundwater Monitoring", T.A. Taylor, G.B.
    Jarvis, H. Xu, A.C. Bevilacqua, and J.E. Kenny, Analytical Instrumentation, 21, 141 (1993.)

9.    "Rapid, subsurface, in-situ  field screening of petroleum hydrocarbon contamination using laser-
    induced fluorescence over optical fibers",  S.  H. Lieberman, G.  A.  Theriault, S. S.  Cooper, P. G.
    Malone, R.  S. Olsen, P. W. Lurk, Proc. Second International Symposium on Field Screening Methods
    for Hazardous Wastes and Toxic Chemicals, Air & Waste Management Association, Pittsburgh, PA,
    1991, 57-63.

10.  "Laser-induced fluorescence in contaminated soils", P. W. Lurk,  S. S.  Cooper, P. G. Malone,  R. S.
    Olsen, S. H. Lieberman, SPIE 1434, 114-118 (1991).

11.  "Remote in-situ determination of fuel products in soil: field results and laboratory investigations", S.
    Apitz, L. M. Borbridge, G. A.  Theriault, S. H. Lieberman, Analusis 20, 461-474 (1992).

12.  "Optimization of the optical characteristics of a fiber-optic guided laser fluorescence technique for
    the in-situ evaluation of fuels in soils",  S. Apitz,  G. A. Theriault,  S. H. Lieberman, SPIE 1637, 241-
    254 (1992).

13.  "The fluorescent response of fuels in soils: insights into fuel-soil interactions", S. Apitz, L. M.
    Borbridge, K. Bracchi, S. H. Lieberman, SPIE 1716 (1992).
                                              87

-------
14. "Subsurface screening of petroleum hydrocarbons in soils via laser induced fluorometry over optical
    fibers with a cone penetrometer system", S. H. Lieberman, S. E.  Apitz, L. M. Borbridge, G.  A.
    Theriault, SPIE 1716 (1992).

15. "Capabilities and limitations of a cone penetrometer deployed fiber optic laser-induced fluorescence
    (LIF) petroleum oil and lubricant (POL)  sensor",  S. H. Lieberman,  W. McGinnis, M. Davey, K.  C.
    Wu, SPIE 2367, 2-16 (1994).

16. "Transportable tunable dye laser for field  analysis  of aromatic hydrocarbons in groundwater," R. W.
    St. Germain, G. D. Gillispie, Proc. Second International Symposium of Field Screening Methods for
    Hazardous  Waste Site Investigations, Las  Vegas, NV, 789-792 (1991).

17. "In-situ Tunable Laser Fluorescence Analysis of Hydrocarbons", R. W. St. Germain, G. D.  Gillispie,
    SPIE 1637, 159-171  (1992).

18. "Variable Wavelength Laser System for field fluorescence measurements," R. W. St. Germain, G.  D.
    Gillispie, and J. L. Klingfus, Proceedings Field Screening Methods for Hazardous Wastes and Toxic
    Chemicals, Air & Waste Management Association, Pittsburgh, PA,  1113-1122 (1993).

19.  "Subsurface optical probes: Current status and future prospects",  in Proceedings Field Screening
    Methods for  Hazardous  Wastes and Toxic Chemicals, Air & Waste  Management Association,
    Pittsburgh, PA, 793-805 (1993).

20.  "Real-time continuous measurement of  subsurface petroleum contamination with the rapid optical
    screening tool (ROST)", R. W. St. Germain and G. D. Gillispie,  in Proceedings Field Screening
    Methods for  Hazardous  Wastes and Toxic Chemicals, Air & Waste  Management Association,
    Pittsburgh, PA, 467-477 (1995).

21.  "Performance Characterization of the rapid optical screening tool (ROST)", G. D. Gillispie and  R.
    W.  St.  Germain, in Proceedings Field Screening Methods for  Hazardous  Wastes  and Toxic
    Chemicals, Air & Waste Management Association, Pittsburgh, PA, 478-489 (1995).

22. "Evaluation of a  Nd:YAG-Pumped Raman  Shifter as a Broad-Spectrum Light Source", George  B.
    Jarvis, Sam Mathew,  and Jonathan E. Kenny, Applied Optics 33, 4938 (1993).

23. * "A Fiber Optic Multichannel Spectrometer System for Remote Fluorescence Detection in  Soils", S.
    J. Hart, Y.-M. Chen, B.  K. Lien, and J. E. Kenny, SPIE 2835, 73, (1996).

24. "Analysis  of Multicomponent Fluorescence Data", I.M. Warner,  G.D.  Christian,  E.R. Davidson,
    Analytical Chemistry, 49, 564  (1977).

25. *  J. Lin, S. J. Hart, T. A. Taylor, J. E. Kenny, "Laser Fluorescence EEM Probe for Cone Penetrometer
    Pollution Analysis", SPIE 2367, 70 (1994).

26. "Cone penetrometer deployed  in situ video microscope for characterizing  sub-surfacesoil prpoerties",
    S.  H. Lieberman, S. S. Knowles,  P. M.  Stang,  J.  Kertesz  and  D. Mendez,  Proceedings Field
    Analytical Methods for Hazardous  Wastes and Toxic Chemicals, Air &  Waste Management
    Association, Pittsburgh, PA, 579-587 (1997).

27. * "Improved Two Fiber Probe  for in situ Spectroscopic Analysis", J.  Lin, S. J. Hart, and J. E. Kenny,
    Anal. Chem., 68, 3098 (1996).

-------
28. "Toxicological Profile for Jet Fuels JP-4 and JP-7", Research Triangle Institute, contract # 205-93-
    0606 and report for U.S. Department of Health and Human Services, Public Health Service, Agency
    for Toxic Substances and Disease Registry (June, 1995).

29. * "Subsurface Contaminant Monitoring by Laser Fluorescence Excitation-Emission Spectroscopy in a
    Cone Penetrometer Probe," J. Lin, S. J. Hart, W. Wang, D. Namychkin and J. E. Kenny, Proc. SPIE
    2504, 59 (1995).

30. * "Field Demonstration of a Multichannel Fiber Optic Laser Induced Fluorescence System in a Cone
    Penetrometer Vehicle," S.  J. Hart, Y.- M.  Chen, J. E. Kenny,  B.  K.  Lien  and T.  W. Best, Field
    Analytical Chemistry and Technology, 1, 343 (1997).

31.  I. B. Berlman, Handbook of Fluorescence Spectra of Aromatic Molecules, 2nd Edition, Academic
    Press Inc., New York, 1971.
                                              89

-------
                       BIBLIOGRAPHY OF OTHER PUBLICATIONS

               RESULTING FROM THIS PROJECT OR RELATED PROJECTS:
Research Monograph:

Laser-Induced Fluorescence for Subsurface  Contaminant Monitoring,  Jonathan E.  Kenny, Jane W.
Pepper, Andrew O. Wright, Yu-Min Chen, Steven L. Schwartz, and Charles G. Shelton, CRC Press, in
press.
Review Article:

"Spectroscopy in  the  Field:  Emerging Techniques for On-Site Environmental Measurements," A.
Henderson-Kinney and J. E. Kenny, Spectroscopy, Vol. 10, No. 7, p. 32 (1995).
Research Articles:

"Two Fiber Spectroscopic Probe with Improved Scattered Light Rejection, A. O. Wright, J. W. Pepper and
J. E. Kenny, Analytical Chemistry, in press.

"Laser-Induced Fluorescence and Fast Gas Chromatography/Mass Spectrometry with Subsurface Thermal
Extraction of Organics: Field Analytical Technologies for Expediting Site Characterization and Cleanup,"
A. Robbat Jr., J. E. Kenny, S. Smarason, J. W. Pepper, and A. O. Wright, Remediation Winter 1998, 95-
111.

"Subsurface  Contaminant Monitoring by  Laser  Excitation-Emission Matrix/Cone  Penetrometer," J.
Pepper, Y.-M. Chen, A. Wright, R. Premasiri, J.E. Kenny, SPIE 3534, pp. 234-242 (1998).

"A Fiber Optic Laser Induced Fluorescence Excitation Emission Detector Applied to Flow Injection
Analysis of PAHs, " S. J. Hart, G. J. Hall, and J. E. Kenny, SPIE 3534, pp. 601-611 (1998).
                                             90

-------
                                           GLOSSARY


Absorption - The interaction of light with a chemical species wherein the molecule absorbs a photon and
makes a transition to a higher energy state. For visible and ultraviolet light, changes in electronic states
are produced.  These are followed by relaxation processes to  lose the excess energy.  These processes can
be non-radiative or radiative, examples of the latter being fluorescence and phosphorescence.

Absorption filter - An optical element that is chosen to absorb a certain range of wavelengths for the
purpose of blocking unwanted light.

BTEX - Class of aromatic compounds consisting of benzene, toluene, ethyl-benzene, and xylenes  (ortho,
meta, and para isomers).

CCD (charged coupled device) -  A semiconductor device consisting of many pixels that can generate and
collect electrons generated by incident light. It is a sensitive  photon counting detector with high signal to
noise ratios and low background used for imaging and multi channel data acquisition.

Counts - The unit  of measurement in a CCD; one count is usually equal to 7-8 photons.

CPT (Cone penetrometer technology) - Technique wherein a steel rod with a tip is pushed hydraulically
into the ground to allow  in situ measurements of many types to be made.  These include geophysical
measurements such as tip pressure, sleeve friction, and conductivity and also chemical measurements such
as LIF, laser induced breakdown spectroscopy, and thermal desorption mass spectrometry.

EEM - Excitation emission matrix,  used to describe three  dimensional fluorescence  data consisting  of
excitation and emission wavelengths and fluorescence intensity.

Emission - The production of light by a molecule due to a transition to a lower energy state.

Excitation - The absorption of light by a molecule due to a transition to a higher energy state.

Extinction coefficient - See molar absorptivity.

Fiber-optics - A light guide that is comprised of a small glass or plastic solid cylinder, called the fiber
core, that  is surrounded by a similar material in contact with it called the cladding.  The refractive index
of the core is greater than the index of the cladding allowing the light to be  totally internally reflected
within the fiber.  This causes light to be propagated from one end of the fiber (launch end) to the opposite
end (distal end).

Fiber-optic connector - An assembly into which a fiber optic is placed so that  it can be mated to another
fiber with as little loss as possible.

Fluorescence - The emission of light due to a transition to a lower energy electronic state of the  same spin
multiplicity by a molecule.

HPLC (high performance  liquid chromatography) - An instrument used to separate molecules in a liquid
flow by passing them over a stationary medium chosen for its chemical properties to retain some species
preferentially, thus enabling a separation.

LIF - (laser induced fluorescence) - The excitation of aromatic molecules  to induce them to fluoresce
using a laser as the light source.
                                                91

-------
LOD  (limit of detection) - The lowest,  meaningful analytical signal used for quantitative analysis.
Commonly defined as 3 times the standard deviation of the baseline signal.

Molar absorptivity - Also called extinction coefficient. Is a measure of a molecule's ability to absorb light
of a certain wavelength.

PAH (polycyclic aromatic hydrocarbon) - An aromatic molecule which is comprised of two or more fused
aromatic rings. Examples include naphthalene, anthracene, and pyrene.

Photon  - A fundamental, quantum unit of light energy that is  equal to Planck's constant times  the
frequency of the light.

Quantum yield (fluorescence) - The number of molecules that emit light as fluorescence divided by the
number of molecules excited by the incident light.

Signal-to-noise ratio - The result of the division of a detector's signal by the signal's standard deviation.

SRS (stimulated Raman scattering) - Process by which high intensity laser light is scattered by molecules
producing light of different wavelengths, both of shorter and longer wavelength than the pump beam.
                                               92

-------
                                             INDEX
absorbtion	16, 29, 87
anthracene	4, 5, 33, 34, 35, 48, 54, 68, 88
aqueous	4, 24, 43
attenuationlO, 11, 23, 27, 28, 29, 41, 56, 57, 59,
  64, 65, 66
aviation gas	43
B
beams ....9, 10, 14, 17, 24, 27, 28, 37, 41, 42, 73
benzene	1, 4, 34, 35, 87
BTEX1, 43, 44, 45, 48, 49, 53, 64, 65, 66, 67,
  87
C
cablelO, 11, 12, 14, 22, 23, 37, 40, 45, 56, 57,
  59
calibrant	23, 27, 28, 29, 46, 47, 57, 59, 71
calibrate	5,22
calibration curves	24
CCD	19, 20, 22, 23, 24, 29, 40, 71, 87
channels, 17, 19, 20, 21, 27, 28, 30, 32, 33, 35,
  38, 47, 48,  54, 55,  57, 59, 61, 64, 65,  66, 67,
  71, 74, 77,  87
concentration, 24, 25, 26, 29, 32, 39, 57, 59,
  66
cone penetrometer technology	1
connectorslO, 12, 13, 18,  19, 22, 23, 24, 37, 45,
  85
CPT1, 4, 7, 8, 11, 13, 21, 23, 37, 38, 39, 40, 41,
  45, 55, 56,  64, 65,  66, 67, 70, 78, 83,  87
cross-talk	20
cyclohexane	4, 25, 26, 27, 32, 33, 34, 35
D
data acquisition	8, 20, 21, 22, 38, 61, 78, 87
depths, 5, 7, 8, 20, 21, 36, 38, 39, 40, 41, 42,
  46, 47, 48,  49, 50,  51, 53, 54, 56, 57,  59, 61,
  62, 64, 66,  67, 73,  74
detectionl, 4, 8, 13, 14, 18, 19, 20, 22, 23, 24,
  27, 41, 49,  56, 57,  59, 74, 87
detector2, 16, 17,  18, 19, 20, 29, 40, 47, 49, 71,
  73, 87, 88
distal cones	14, 24
EEM2, 3, 5, 7, 8, 22, 29, 32, 33, 34, 35, 38, 46,
  49, 50, 51, 53, 64, 65, 67, 68, 74, 76, 77, 84,
  85,87
emission2, 4, 17, 23, 27, 28, 29, 33, 35, 38, 41,
  49, 54, 64, 65, 68, 73, 74, 76, 77, 87
EPA	39, 45, 55, 71
excitationl, 2, 4, 5, 9,  10, 11, 14, 16, 17, 18, 22,
  23, 26, 27, 28, 29, 30, 31, 32, 33, 35, 36, 37,
  38, 41, 42, 46, 47, 48, 49, 54, 59, 62, 64, 65,
  66, 68, 71, 73, 76, 77, 87
  48
fiberl, 3, 9, 10, 11, 12, 14, 15, 16, 17, 18, 19,
  20, 21, 22, 23, 24, 26, 27, 28, 29, 30, 37, 39,
  47, 55, 56, 57, 59, 64, 65, 66, 71, 78, 84, 85,
  87
fiber bundle	3,23
fiber cable	10, 16, 23, 37
fiber optics	1, 23, 30, 47
fibersl, 10, 11, 12, 13, 14, 15, 16, 18, 19, 20,
  22, 37, 40, 56, 57, 59, 73
field data. 2, 3, 5, 7, 8, 25, 26, 33, 38, 39, 49, 59
  37
filters	14, 16, 17, 18, 19, 28, 37, 41
fluorescence2, 3, 5, 8,  13, 14, 15, 16, 17,  20, 21,
  22, 23, 24, 25, 26, 27, 28, 29, 30, 31, 32, 33,
  34, 35, 38, 39, 40, 41, 46, 47, 48, 49, 50, 51,
  53, 54, 57, 59, 61, 62, 64, 66, 67, 68, 73, 74,
  77, 84, 85, 87, 88
fluorescence intensity	2, 3, 32, 54, 87
fluorescence response	16, 30, 46, 47
  31
focal plane	10,22
  4, 5, 24, 25, 26, 32,  34, 39, 40, 41, 43,  44, 48,
  49, 51, 53, 54, 56, 57, 59, 62, 64, 65, 66, 68,
  71,76
G
groundwater	7, 14, 43, 44, 49, 51, 84, 86
H
Hanscom AFB5, 7, 8, 32, 42, 43, 44, 45, 46, 47,
  57, 59, 61, 70, 76
HPLC	7, 8, 22, 83, 87
hydraulic	1, 12, 18, 21, 22, 37
improvement factors	15
in situl, 2, 5, 22, 25, 31, 46, 49, 53, 54, 59, 64,
  65, 67, 68, 71, 73, 76, 83, 84, 85, 87
Incident energy	28
instrument!, 2, 4, 7, 8, 21, 25, 26, 37, 38, 41,
  45, 46, 47, 48, 49, 55, 57, 59, 70, 71, 87
                                               93

-------
intensity
                             2, 22, 50, 51, 88
jet fuel ...5, 25, 32, 33, 34, 48, 49, 51, 54, 66, 77
L
laserl, 3, 9, 22, 23, 30, 37, 40, 41, 42, 45, 47,
  56, 57, 59, 71, 83, 84, 86, 87, 88
least-squares [[[ 5
LIF1, 2, 4, 7, 8, 9, 20, 21, 22, 24, 25, 26, 29, 33,
  35, 37, 38, 39, 40, 41, 45, 56, 57, 59, 61, 64,
  65, 66, 67, 70, 71, 78, 83, 87
LIF-EEM1, 2, 7, 8, 21, 22, 24, 25, 26, 29, 78,
  83
lightl, 2, 10, 11, 13,  14,  16, 17,  18, 19, 20,  22,
  23, 27, 28, 29, 31, 36, 37, 41, 42, 43, 47, 48,
  50, 51, 54, 56, 57, 59, 73, 84, 87, 88
limit of detection .................................. See  LOD
LNAPL [[[ 43
LOD .................................................. 24, 27, 87
M
mercury vapor lamp ....................................... 22
mirror [[[  9, 14
molar absorptivity ............................................ 2
monitoring ......................................... 48, 49, 76
TV
N.A ................................ See numerical aperture
naphthalene4, 5, 33,  34,  35, 48,  51, 54, 64,  65,
  67, 74, 76, 77, 88
neural networks ............................................... 5
numerical aperture ......................................... 19
O
optical breadboard ................................ 9, 21, 37
optical fibers ......................... 1,  9, 14, 41, 83, 86
PAH
pattern
pattern recognition
phenol
                                        1,88
                          5, 32, 33, 54, 68, 77
                                           5
                          4, 15, 24, 33, 35, 36
pixels ................................................. 20, 24, 87
plume ............... 7, 48, 49, 50, 51, 57, 59, 62, 64
power ............... 27, 29, 30, 31, 37, 38, 41, 47, 55
power normalization ................................ 27, 38
prism [[[ 9
probel,  3, 4, 5, 7, 13, 14, 15, 16,  18, 20, 22, 24,
  25, 26, 27, 28, 29, 30, 33, 37, 38, 39, 40, 45,
  47, 56, 64, 65, 67, 71, 73, 78, 83, 84, 85
pulse energies	10
Q
quantum yield	2, 29
quinine sulfateS, 27, 28, 29, 30, 31, 37, 38, 46,
  47,71
Quinine sulfate	27, 32, 71
R
Raman	1, 9, 10,  11, 17, 23, 37, 73, 84, 88
Raman shifter	9, 73
rank annihilation	2, 5
ray traces	19
Rayleigh scattering	16, 17, 41
reference EEMs	5, 53, 77
rod	1, 11, 13, 14, 41, 87
                                                    sand	4, 24, 25, 26, 27, 32, 35, 36, 71, 76
                                                    sapphire window	1, 13, 14, 15,66
                                                    saturated zone	25, 26
                                                    sensors	7
                                                    site characterization	2, 46, 48, 70, 71, 73
                                                    Software	7
                                                    soil	43
                                                    solution4, 5, 10, 15, 18, 25, 26, 27, 28, 35, 37,
                                                      46, 47, 71
                                                    solvent	24
                                                    spectrometer	1
                                                    SRS	9,73,88
                                                    subsurface	1, 3, 71
transmission	17
V
volatiles	1
W
water	17, 24, 33, 41, 44, 49, 54, 57, 59

-------
         299 nm
         \
          257 nm
    -C
    +J
    Q<
    4>
    Q
315 nm
 278 nm
 327 nm
                                   -2 E-08  1 E-OS   4 E-OE
266 nm
                                                    342 nm
                                                 0 E+OO   2 E-OS   4 E-0£
                                                   289 nm
              Summed fluorescence (relative units)

Figure 1. Depth vs. summed fluorescence plots for each
 excitation wavelength during push 2.
                                                                           CD
                                                                           P
                                                                           P
                                                                           H^
                                                                           X

-------
       299 nm
315 nm
327 nm
    0 E+00   5 E-08    1 E-07
       257 nm

               266 nm
                                -1 E-07   1 E-07   3 E-07
                                                  342 nm
                                                 289 nm
                                               0 E+OO  2 E-07  4 E-07
            Summed fluorescence (relative units)
Figure 2. Depth vs. summed fluorescence plots for each
 excitation wavelength during push 3.

-------
         299 nm
315 nm
327 nm
                                                    342 nm
                      I
                        228jiJii_
      -3 E-01   8 E-01   2 E+00   -5 E-02  5 E-02  2 E-01
               266 nm
                             289 nm
               Summed fluorescence (relative units)
Figure 3. Depth vs. summed fluorescence plots for each
 excitation wavelength during push 4.

-------
         299 nm
315 nm
327 nm
         257 nm
278 nm
                                                 342 nm
                    0 E+OO 1 E-Q9  2 E-09  3 E-09
266 nm
                    0 E+OO  5 E-02  1 E-01  2 E-01
                                 -5 E-OS 0 E+OO 5 E-08 1 E-07 2 E-i
                                              0 E+OO   1 E-09   2 E-09
             289 nm
             Summed fluorescence (relative units)
Figure 4. Depth vs. summed fluorescence plots for each
 excitation wavelength during push 5.

-------
       299 nm
315 nm
327 nm
                                                    342 nm
       257 nm
278 nm
266 nm
                                                     289 nm
   -2 E-01 3 E-01 8 E-01 1 E+00 2 E+00    0 E+00 5 E-01 1 E+00  2 E+00    ° E+0°   9 E"07   2 E"06
                                                 -2 E-07 0 E+00 2 E-07 4 E-07 6 E-i
            Summed fluorescence (relative units)
Figure 5. Depth vs. summed fluorescence plots for each
 excitation wavelength during push 6.

-------
          299 nm
         257 nm
        t
315 nm
      0 E+00 2 E-07 4 E-07 6 E-07 8 E-07   Q E+00   1 E-08   2 E-08
 278 nm
327 nm
                                     0 E+00  1 E-07   2 E-07
                                      266 nm
                                      «m"""
342 nm
                                                      •*•"	--
                             289 nm
               Summed fluorescence (relative units)
Figure 6. Depth vs. summed fluorescence plots for each
 excitation wavelength during push 7.

-------
        299 nm
315 nm
 327 nm
        257 nm
278 nm
                                                  342 nm
                                 -3 E-09 2 E-09 7 E-09 1 E-OS
266 nm
              289 nm
              Summed fluorescence (relative units)
Figure 7. Depth vs. summed fluorescence plots for each
 excitation wavelength during push 8.

-------
      299 nm
315 nm
327 nm
                                                342 nm
      257 nm
 ^—•
 Q<
 4>

 Q
278 nm
266 nm
                 -5 E-02    2 E-01   4 E-01
              289 nm
                                              0 E+00 1 E-08 2 E-08 3 E-08 4 E-08
            Summed fluorescence (relative units)

Figure 8. Depth vs. summed fluorescence plots for each

 excitation wavelength during push 9.

-------
      299 nm
315 nm
   327 nm
      257 nm
278 nm
 &
 JS
0 E+00 7 E-09 1 E-08 2 E-08


   266 nm
                                                342 nm
                 289 nm
              Summed fluorescence (relative units)

Figure 9. Depth vs. summed fluorescence plots for each

 excitation wavelength during push 10.

-------
         299 nm
       257 nm
315 nm
     0 E+OO 1 E-07  2 E-07  3 E-07   0 E+OO 1 E-08 2 E-08 3 E-08
278 nm
327 nm
                                                     342 nm
                                                  -3 E-08   7 E-08  2 E-07
266 nm
              289 nm
                                                 0 E+OO 1 E-06 2 E-06 3 E-06 4 E-06
               Summed fluorescence (relative units)
Figure 10. Depth vs. summed fluorescence plots for each
 excitation wavelength during push 11.

-------
        299 nm
        257 nm
&
JS
                    315 nm
                    278 nm
     -3 E-01  3 E-01 8 E-01 1 E+OO     -1 E-01  4 E-01  9 E-01 1 E+OO
327 nm
266 nm
                                      V
342 nm
                                                  0 E+OO 4 E-09  S E-09 1 E-Oi
289 nm
                                                  0 E+OO 1 E-08 2 E-08 3 E-08 4 E-08
               Summed fluorescence (relative units)

Figure 11. Depth vs. summed fluorescence plots for each

 excitation wavelength during push 12.

-------
         299 nm
         257 nm
  &
  JS
315 nm
     -1 E-06 0 E+00 1 E-06 2 E-06 3 E-06    -1 E-09   9 E-09   2 E-Oi
278 nm
     -3 E-01 3 E-01 8 E-01 1 E+00     -2 E-01 3 E-01 8 E-01 1 E+00
327 nm
266 nm
                                     -9 E-07  0 E+00 9 E-07  2 E-06
                                                        342 nm
                                                     0 E+00   2 E-08   4 E-OE
289 nm
                                                     0 E+00 5 E-09 1 E-08 2 E-08
                Summed fluorescence (relative units)

Figure 12. Depth vs. summed fluorescence plots for each

 excitation wavelength during push 13.

-------
        299 nm
        \
        257 nm
315 nm
278 nm
327 nm
266 nm
                                                  342 nm
              289 nm
                                                 \
                                               0 E+00 1 E-08 2 E-OS 3 E-08 4 E-Oi
              Summed fluorescence (relative units)
Figure 13. Depth vs. summed fluorescence plots for each
 excitation wavelength during push 14.

-------
                     299 nm
                       315 nm
327 nm
                                                  342 nm
                0 E+OO 2 E-07 4 E-07  6 E-Oi
                    257 nm
g,
JS
                               0 E+OO  3 E-08 6 E-08 9 E-Oi
                       266 nm
278 nm
                                                            0 E+OO  2 E-OS  4 E-08  6 E-Oi
289 nm
                                                                                       CD
                                                                                       X
                                                                                       Qd
                         Summed fluorescence (relative units)
Figure 1. Depth vs. summed fluorescence for each excitation wavelength for push 1.

-------
                       299 nm
 315 nm
 327 nm
                 0 E+OO  3 E-06  6 E-06  9 E-06
                     257 nm
                 0 E+OO  1 E+OO  2 E+OO 4 E+OO
                                0 E+OO 3 E-08 6 E-08 9 E-08
266 nm
                                                               342 nm
278 nm
                                                             0 E+OO 2 E-08  4 E-08 7 E-Oi
289 nm
18
16 •
12 •
10 •
8 •
6 •
0 E


K~ — •— — «

-------
                      299 nm
 315 nm
 327 nm
                   257 nm
                0 E+00 1 E+00 2 E+00  4 E+00
266 nm
278 nm
                                                            342 nm
289 nm
                        Summed fluorescence (relative units)
Figure 3. Depth vs. summed fluorescence for each excitation wavelength for push 3.

-------
                    299 nm
315 nm
327 nm
                    257 nm
             -C
             +J

             Q<
             4>

             Q
                0 E+00 1 E+00 2 E+00  4 E+00
266 nm
                               0 E+00 2 E-06 3 E-06  5 E-06
278 nm

18 •
16 •
12 •
10 •
6 •
4 •


X.
J=L.

0 E+00 1 E+00 2 E+00 3 E+00
                                                              342 nm
289 nm

18 •
16 •
12 |
10 •
6 •
4 •



^
r~

0 E+00 2 E-05 4 E-05 6 E-05
                         Summed fluorescence (relative units)


Figure 4. Depth vs. summed fluorescence for each excitation wavelength for push 4.

-------
                        299 nm
 315 nm
 327 nm
                     257 nm
266 nm
                 -2 E-01 3 E-01 8 E-01 1 E+00 2 E+00    -1 E-06  1 E-06  3 E-06 5 E-Of
278 nm
                                                                342 nm
289 nm
                                                E+00  1 E+00  2 E+00 3 E+00   0 E+00  2 E-06  4 E-06 6 E-Of
                          Summed fluorescence (relative units)
Figure 5. Depth vs. summed fluorescence for each excitation wavelength for push 5.

-------
                       299 nm
 315 nm
 327 nm

16 •
12 i
10 •
6 •
0 E

L_
^
00 5 E-06 1 E-05 2 E-05
                     257 nm
18
16'
12
10 •
6 •


k
X
._-— --^
\
0 E+00 1 E+00 2 E+00 4 E+00
                  E+00  5 E-06  1 E-05  2 E-05   0 E+00 1 E-07  2 E-07  3 E-07    0 E+00 2 E-07 3 E-07 5 E-07     0 E+00 1 E-08  2 E-08 3 E-l
                                                               342 nm
266 nm
278 nm

16 -
12 -
10 -
6 -

^
£•'
0 E+00 1 E-08 2 E-08 3 E-08
289 nm
18 •
16 •
12 •
10 •
6 •



t
^
0 E+00 1 E-05 2 E-05 3 E-05
                          Summed fluorescence (relative units)
Figure 6. Depth vs. summed fluorescence for each excitation wavelength for push 6.

-------
                        299 nm
 315 nm
 327 nm
             -C
             +J

              Q<
              4>

             Q
18 ,
16 •
12 •
io ;
6 •
0 •
0 E



J
^___ZZI-— ^-":"n 	 'HH
00 1 E-06 2 E-06 3 E-06
                     257 nm
                  0 E+00 6 E-01  1 E+00 2 E+00
                                  0 E+00  1 E-07 2 E-07  3 E-0
266 nm
16 •
14 •
12 •
10 •
8 •
6 •



r
>•
                                                                 342 nm
278 nm
                                  0 E+00  6 E-07  1 E-06  2 E-06     0 E+00 1 E-01  2 E-01 4 E-i
                                                               0 E+00  5 E-09  1 E-08 2 E-Oi
289 nm
                          Summed fluorescence (relative units)


Figure 7. Depth vs. summed fluorescence for each excitation wavelength for push 7.

-------
                        299 nm
 315 nm
 327 nm
                     257 nm
             -C
             +J

              Q<
              4>

             Q
                                 0 E+00 6 E-OS  1 E-07  2 E-0
266 nm
                                                                 342 nm
278 nm
                 0 E+00 S E-01  2 E+00  2 E+00     0 E+00 3 E-06  6 E-06  9 E-06   0 E+00  2 E+00  4 E+00 6 E+00   0 E+00  5 E-06  1 E-05  2 E-05
289 nm
24 •
20 •
18 •
16 •
14 •
12 •
10 •
6 •
2 •
0 •
0 E





b?
f
Se~
^
^Tfen 	 , ,
______ il y* i '
X

00 5 E-06 1 E-05 2 E-05
                           Summed fluorescence (relative units)


Figure 8. Depth vs. summed fluorescence for each excitation wavelength for push 8.

-------
                        299 nm
 315 nm
 327 nm
                     257 nm
             -C
             +J

             Q<
             4>

             Q
                 0 E+00  8 E-01  2 E+00 2 E+00
266 nm
                                0 E+00  3 E-06  6 E-06 9 E-06
                                                                342 nm
278 nm
                                               0 E+00 8 E-01  2 E+00 2 E+00
                                                              0 E+00  7 E-09  1 E-08 2 E-i
289 nm
                          Summed fluorescence (relative units)


Figure 9. Depth vs. summed fluorescence for each excitation wavelength for push 9.

-------
                       299 nm
 315 nm
 327 nm
                     257 nm
             -C
             +J

             Q<
             4>

             Q

16 •
12 •
8 •
6 •
t
f
1
1

h
i
i
H
H
i
•1
1
H
H
L
t
1
h»H
0 E+00 6 E-01 1 E+00 2 E+00
266 nm
18 -,
16 .
14 .
12 .
10 •
6 •





\
0 E+00 6 E-07 1 E-06 2 E-06
                                                                342 nm
278 nm

16 •
14 •
12 •
10 •
6 •
I
f
\
h i
H H
T"1
0 E+00 6 E-02 1 E-01 2 E-01
289 nm
16 •
14 •
10 •
6 •

H
H
t
H
h
H
h
1-
K
0 E+00
1
1
i
*L
^
3r;
T1
lie
x
^
X
y
HH
2 E-07 4 E-07 6 E-07
                  0 E+00  6 E-01  1 E+00 2 E+00    0 E+00 6 E-07 1 E-06 2 E-06    0 E+00 6 E-02  1 E-01  2 E-01     0 E+00  2 E-07 4 E-07  6 E-i



                          Summed fluorescence (relative units)



Figure 10. Depth vs. summed fluorescence for each excitation wavelength for push 10.

-------
                         299 nm
 315 nm
 327 nm
              -C
              +J

               Q<
               4>

              Q
                      257 nm
12 •
10 •
6 •
4 •



1
fcp-^ .
^ •
0 E+00 2 E-07 3 E-07 5 E-07
                   E+OO 1 E-06  3 E-06  4 E-06    0 E+00 2 E-07 3 E-07  5 E-07     0 E+00 3 E-07  6 E-07 9 E-07   0 E+00  6 E-OS  1 E-07  2 E-07
                                                                     342 nm
10 -
8 -
4 -



*— — — ^^ 	 », ,
»!_—-— — — ^^~* '
^TlM
l-teL
JE?#— i
h^q^
Oq
A
0 E+00 6 E-OS 1 E-07 2 E-07
266 nm
278 nm
289 nm
                  -4 E-01 2 E-01 8 E-01 1 E+00 2 E+00     0 E+00 1 E-06  3 E-06 4 E-06    0 E+00 7 E-01 1 E+00 2 E+01
                                                                   0 E+00 2 E-06 4 E-06 6 E-06
                            Summed fluorescence (relative units)


Figure 11. Depth vs. summed fluorescence for each excitation wavelength for push 11,

-------
                         299 nm
 315 nm
 327 nm
                   0 E+00  5 E-06  1 E-05  2 E-05
                      257 nm
                   0 E+00 1 E+00 2 E+00 3 E+00 4 E+00
                                   E+00  1 E-07  2 E-07 3 E-07
266 nm
                                   0 E+00 2 E-06  3 E-06 5 E-06
                                                                   342 nm
                                                                  0 E+00 9 E-09 2 E-08 3 E-0
278 nm
289 nm
                                                   0 E+00  9 E-01  2 E+00 3 E+00
                                                                  0 E+00  6 E-06  1 E-05  2 E-05
                           Summed fluorescence (relative units)
Figure 12. Depth vs. summed fluorescence for each excitation wavelength for push 12.

-------
                      299 nm
 315 nm
 327 nm
                                                           342 nm
                                                          1
                0 E+OO 6 E-08  1 E-07  2 E-0'i
                   257 nm
266 nm
278 nm
289 nm



£
JS
&4
Q




16 •
12 •
10 •

8.
6 •
4 •

t







1
H
1














0 E+OO 6 E-01 1 E+OO 2 E+OO











16 -
12-
10 -

8-
6 -
4 -

f
•




%









0 E+OO 6 E-07 1 E-06 2 E-06











16 •
12-
10 •

8 •
6 •
4-

1
1
1
i
'
H

!
!
A ,








0 E+OO 2 E-01 3 E-01 5 E-01










18
16 •
12 •
10 •

8 •
6 •
4


1





1
1
1
1




fc
h|,
T
T.
T
*\
HH
0 E+OO 3 E-07 6 E-07 9 E-07
                        Summed fluorescence (relative units)
Figure 13. Depth vs. summed fluorescence for each excitation wavelength for push 13.

-------
                        299 nm
 315 nm
 327 nm
                   0 E+00 2 E-07 3 E-07 5 E-07
                      257 nm
                                  J
266 nm
                                                                  342 nm
278 nm
                  0 E+00 1 E+00 2 E+00 3 E+00 4 E+00    0 E+00 6 E-07  1 E-06  2 E-06     0 E+00  2 E-01 3 E-01  5 E-01    0 E+00 2 E-06 3 E-06 5 E-06

16 •
12 •
10 •
6 •
H
K
h
h
h
h

H
H
H
-\
S^
3z
^r
Jz"1
He
£
±
^
0 E+00 2 E-01 3 E-01 5 E-01
                                                                  I	8	1
                                                                 0 E+00  1 E-08 2 E-08 3 E-OS
289 nm

16 -
12 -
10 -
I
6 •
1
H
H
H
H
H
L ,
- —i
- H
- H
-« ~\
- -\
	 1
- H
- H
0 E+00 2 E-06 3 E-06 5 E-06
                           Summed fluorescence (relative units)
Figure 14. Depth vs. summed fluorescence for each excitation wavelength for push 14.

-------
                         299 nm
 315 nm
 327 nm
                      257 nm
                  0 E+00 1 E+00 2 E+00 4 E+00
                   E+00  5 E-Q6  1 E-05  2 E-05    0 E+00  6 E-08  1 E-07 2 E-07    0 E+00 2 E-07  3 E-07  5 E-07     0 E+00  1 E-08  2 E-08  3 E-l
                                                                    342 nm
266 nm
278 nm
                                   0 E+00  1 E-06  2 E-06 4 E-06    0 E+00  1 E+00  2 E+00  4 E+00    0 E+00 6 E-06  1 E-05  2 E-05
16 -
10 -
6 -
0 -
0 E

L.
>=~

00 1 E-08 2 E-08 3 E-08
289 nm
                            Summed fluorescence (relative units)
Figure 15. Depth vs. summed fluorescence for each excitation wavelength for push 15.

-------
                      299 nm
 315 nm
 327 nm
                  0 E+OO  1 E-06 2 E-06  3 E-06
                      257 nm
                  0-. I
                                  0 E+OO 3 E-08 6 E-OS 9 E-08
266 nm
                                                                    342 nm
278 nm
                  0 E+OO  S E-01 2 E+OO  2 E+OO     0 E+OO  2 E-06  4 E-06  6 E-06    0 E+OO  6 E-01  1 E+OO 2 E+OO    0 E+OO  1 E-05 2 E-05  3 E-05
                                                                 0 E+OO 2 E-09 4 E-09 7 E-09
289 nm
18 •
16 •
12 •
10 •
6 •
0 •
0 E




V 	 ^^
«_—•»- — "^
3
00 1 E-05 2 E-05 3 E-05
                           Summed fluorescence (relative units)
Figure 16. Depth vs. summed fluorescence for each excitation wavelength for push 16.

-------
                     299 nm
315 nm
327 nm
                     257 nm
             -C
             +J

              Q<
              4>

             Q
                 0 E+OO  S E-01 2 E+00 2 E+00
266 nm
                                 0 E+00  2 E-06  4 E-06  6 E-06
                             342 nm
278 nm
                                                0 E+00  5 E-0
                                                              0 E+00 2 E-09  3 E-09  5 E-09
289 nm
                                                               0 E+00  2 E-06 4 E-06 6 E-06
                          Summed fluorescence (relative units)


Figure 17. Depth vs. summed fluorescence for each excitation wavelength for push 17.

-------
                     299 nm
315 nm
327 nm
                 0 E+OO 1 E-06 2 E-06 4 E-06
                     257 nm
                 0 E+OO  3 E-01  6 E-01  9 E-0'
                                0 E+OO 1 E-08  2 E-08  3 E-08
266 nm
                            342 nm
278 nm
289 nm
                                                              0 E+OO  1 E-06 2 E-06  3 E-06
                          Summed fluorescence (relative units)
Figure 18. Depth vs. summed fluorescence for each excitation wavelength for push 18.

-------
                      299 nm
315 nm
327 nm
                   0 E+OO 7 E-07  1 E-06  2 E-06    0 E+OO 3 E-OS 6 E-08 9 E-0!
                     257 nm
                  0 E+OO  6 E-01  1 E+OO 2 E+OO
266 nm
                             342 nm
278 nm
                                  E+OO 1 E-06 2 E-06 4 E-06    0 E+OO 2 E-01 3 E-01 5 E-0
                                                                0 E+OO 1 E-09 2 E-09 4 E-09
289 nm
                           Summed fluorescence (relative units)
Figure 19. Depth vs. summed fluorescence for each excitation wavelength for push 19.

-------
                      299 nm
315 nm
327 nm
                      257 nm
                  0 E+OO 3 E-01 6 E-01 9 E-0
                   E+00  6 E-07  1 E-06 2 E-06    0 E+OO  5 E-09  1 E-08  2 E-08   0 E+OO  2 E-08  4 E-08  6 E-08   0 E+OO  2 E-09  4 E-09 6 E-09
                              342 nm
266 nm
278 nm
                                  -3 E-07 5 E-07 1 E-06 2 E-06
                                                  0 E+OO  1 E-01  2 E-01  4 E-01
289 nm
                                                                  0 E+OO 1 E-06 2 E-06 3 E-06
                            Summed fluorescence (relative units)
Figure 20. Depth vs. summed fluorescence for each excitation wavelength for push 20.

-------
                     299 nm
                     257 nm
             -C
             +J

              Q<
              4>

             Q
                 0 E+OO 3 E-01 6 E-01 9 E-01
315 nm
                                 0 E+OO 1 E-08  2 E-08  3 E-08
266 nm
                                 0 E+OO 8 E-07 2 E-06  2 E-06
327 nm
                             342 nm
                                                              0 E+OO 2 E-09 3 E-09 5 E-09
278 nm
289 nm
                                                               0 E+OO  1 E-06  2 E-06  3 E-06
                          Summed fluorescence (relative units)


Figure 21. Depth vs. summed fluorescence for each excitation wavelength for push 21,

-------
                     299 nm
315 nm
327 nm

16 •
12 •
10 •
6 •
0 E

>~
}
00 1 E-06 2 E-06 4 E-06
                     257 nm
                 0 E+OO  3 E-01  6 E-01 9 E-01

16 •
12
10 •
6 •
0 E

^
Tr~"^
±
?
T
+00 1 E-08 2 E-08 3 E-08
266 nm
                                0 E+OO 8 E-07 2 E-06 2 E-06
                            342 nm
278 nm
                                               0 E+OO 2 E-01 4 E-01 6 E-01
289 nm
                                                              0 E+OO  1 E-06  2 E-06 3 E-06
                          Summed fluorescence (relative units)
Figure 22. Depth vs. summed fluorescence for each excitation wavelength for push 22.

-------
                     299 nm
                    257 nm
             €H

             f
                0 E+00 2 E-01 3 E-01 5 E-01
315 nm
327 nm
266 nm
278 nm
                                0 E+00  3 E-07  5 E-07  8 E-07
                            342 nm

18
16
14
10
8
6
4
2



•






i
f
1








0 E+00 9 E-09 2 E-08 3 E-08











18 •
16 •
14 •
10 •
8 •
6 •
4
2 •

1
i
s




)
H
1


1
1
1
1








0 E+00 2 E-09 3 E-09 5 E-09











18 •
16 •
14 •
10 •
,.
6 •
4 •

0 E

i

.


1




(
\








+00 3 E-09 6 E-09 9 E-09











18 •
16 •
14 •
10 •
,.
6 •
4
2-

H
h
h
h
h
H
h
1-
K
h
h
h
h
H
H
1
H
H
H
t1
H
-1
H
-\
t
0 E+00 3 E-10 6 E-10 9 E-10
289 nm
                                                             0 E+00 6 E-08 1 E-07 2 E-07
                         Summed fluorescence (relative units)
Figure 23. Depth vs. summed fluorescence for each excitation wavelength for push 23.

-------
                   299 nm
                   257 nm
            -C
            +j

            Q<
            4>

            Q
315 nm
266 nm
327 nm
278 nm
                          342 nm
289 nm
                        Summed fluorescence (relative units)

Figure 24. Depth vs. summed fluorescence for each excitation wavelength for push 24.

-------
                     299 nm
315 nm
327 nm
                            342 nm
                 0 E+00 3 E-OS 6 E-08 9 E-OS
                                                            0 E+00 8 E-10 2 E-09 2 E-09
                    257 nm
             -C
             +J




             Q
266 nm
16 •
12 •
10 •
6 •


i







• >

}
0 E+00 3 E-07 5 E-07 8 E-07
278 nm
18
16 •
12 •
10 •
6 •
h
1-
H
h
h
1-
H
H
1-
h
H
H
H
H
-\
H
-1
-1
i
0 E+00 2 E-02 3 E-02 5 E-02
289 nm
                                                             0 E+00 1 E-07  2 E-07  4 E-07
                         Summed fluorescence (relative units)

Figure 25. Depth vs. summed fluorescence for each excitation wavelength for push 25.

-------
                    299 nm
315 nm
327 nm
                           342 nm
18
16
10
8
6


h
h
1-
1-
1-
1-
l->
1-
H
1-
h-
1-
H
^
H
H
H
H
-1
H
H
H
h
t
0 E+00 1 E-08 2 E-08 4 E-08
                    257 nm
             -C
             +J




             Q
                0 E+00 2 E-01  3 E-01  5 E-01
266 nm
18
16 •
12 •
10 •
6 •

j
H
M
h
H
h
1
H
H
-\
-\
H
H
X
1— 1— I
0 E+00 3 E-07 5 E-07 8 E-07
278 nm

                                                           0 E+00 8 E-10 2 E-09 2 E-09
289 nm
18 •
16 •
12 •
10 •
6 •
0 •

m
\
1 1
i
1 1
t \
t \
\ i
* i
1
h*H
X
00 1 E-07 2 E-07 4 E-07
                         Summed fluorescence (relative units)

Figure 26. Depth vs. summed fluorescence for each excitation wavelength for push 26.

-------
                   299 nm
315 nm
327 nm
                          342 nm
                   257 nm
            -C
            +J

            Q<
            4>

            Q
266 nm
278 nm
289 nm
                        Summed fluorescence (relative units)


Figure 27. Depth vs. summed fluorescence for each excitation wavelength for push 27.

-------
    Lab Analysis of
     Core samples
             LIF signal at each excitation A,
                   measured in situ
Naphthalenes  Total BTEX    257 nm
                                     266 nm
   278 nm
289 nm
299 nm
315 nm
327 nm
342 nm
S^
JS
                                   T
    Concentration (nig/Kg)
Summed fluorescence (relative units)
      Figure 1. A comparison of laboratory core sample analysis (32CPT-2) with in situ CPT-LIF
      measurements at each excitation wavelength for CPT push #2.
                                                                                                            ft
                                                                                                            S3

-------
    Lab Analysis of
     Core samples
                     LIF signal at each excitation A,
                           measured in situ
    Naphthalenes Total BTEX    257 nm
266 nm
278 nm
289 nm    299 nm
315 nm
-*^

Q
                                                                                "7
327 nm
342 nm

    Concentration (nig/Kg)
        Summed fluorescence (relative units)
     Figure 2. A comparison of laboratory core sample analysis (32CPT-2) with in situ CPT-LIF
     measurements at each excitation wavelength for CPT push #3.

-------
    Lab Analysis of
     Core samples
                     LIF signal at each excitation A,
                           measured in situ
    Naphthalenes Total BTEX    257 nm
266 nm
278 nm
289 nm
299 nm
315 nm
327 nm
342 nm
f
Q
    Concentration (mg/Kg)
        Summed fluorescence (relative units)
     Figure 3. A comparison of laboratory core sample analysis (32CPT-2) with in situ CPT-LIF
     measurements at each excitation wavelength for CPT push #4.

-------
    Lab Analysis of
     Core samples
                     LIF signal at each excitation A,
                           measured in situ
    Naphthalenes Total BTEX    257 nm
266 nm
278 nm
289 nm
299 nm
315 nm
327 nm
342 nm
f
Q
    Concentration (nig/Kg)
        Summed fluorescence (relative units)
     Figure 4. A comparison of laboratory core sample analysis (32CPT-1) with in situ CPT-LIF
     measurements at each excitation wavelength for CPT push #5.

-------
Lab Analysis of
 Core samples
                                                          LIF signal at each excitation A,
                                                                measured in situ
Naphthalenes  Total BTEX     257 nm
                                     266 nm
278 nm
289 nm
299 nm
315 nm
327 nm
342 nm
f
Q
                                                                                        ^t—	
                                                                                        .•Big^   I
Concentration (nig/Kg)
                                             Summed fluorescence (relative units)
  Figure 5. A comparison of laboratory core sample analysis (32CPT-3) with in situ CPT-LIF
  measurements at each excitation wavelength for CPT push #6.

-------
Lab Analysis of
 Core samples
                      LIF signal at each excitation A,
                            measured in situ
Naphthalenes  Total BTEX     257 nm
 266 nm
                                                278 nm
289 nm
299 nm
315 nm
327 nm
342 nm
a
0)
Q
                     r
T
Concentration (mg/Kg)
         Summed fluorescence (relative units)
  Figure 6. A comparison of laboratory core sample analysis (32CPT-7) with in situ CPT-LIF
  measurements at each excitation wavelength for CPT push #7.

-------
Lab Analysis of
 Core samples
                     LIF signal at each excitation A,
                           measured in situ
Naphthalenes  Total BTEX    257 nm
266 nm
278 nm
289 nm    299 nm
315 nm
Concentration (nig/Kg)
        Summed fluorescence (relative units)
327 nm
342 nm
  Figure 7. A comparison of laboratory core sample analysis (32CPT-4) with in situ CPT-LIF
  measurements at each excitation wavelength for CPT push #9.

-------
    Lab Analysis of
     Core samples
                     LIF signal at each excitation A,
                           measured in situ
    Naphthalenes Total BTEX    257 nm
266 nm
278 nm
f
Q
    Concentration (mg/Kg)
289 nm
299 nm
315 nm
327 nm
342 nm
        Summed fluorescence (relative units)
     Figure 8. A comparison of laboratory core sample analysis (32CPT-6) with in situ CPT-LIF
     measurements at each excitation wavelength for CPT push #10.

-------
Lab Analysis of
 Core samples
                                      LIF signal at each excitation A,
                                           measured in situ
      TPH
257 nm
£
-=
    Cone 
-------
Lab Analysis of
 Core samples
                                      LIF signal at each excitation A,
                                           measured in situ
      TPH
257 nm      266 nm     278 nm      289 nm     299 nm      315 nm      327 nm     342 nm
f
Q
    Cone 
-------
Lab Analysis of
 Core samples
                                      LIF signal at each excitation A,
                                           measured in situ
      TPH
257 nm
f
Q
266 nm
278 nm
289 nm    299 nm
315 nm
327 nm    342 nm
    Cone 
-------
Lab Analysis of
 Core samples
                           LIF signal at each excitation A,
                                measured in situ
     TPH
   Cone
               257 nm
266 nm
278 nm
289 nm
299 nm
315 nm
327 nm
342 nm
        Summed fluorescence (relative units)
  Figure 12. A comparison of laboratory core sample analysis (32CPT-1) with in situ CPT-LIF
  measurements at each excitation wavelength for CPT push #5.

-------
 Lab Analysis of
  Core samples
                   LIF signal at each excitation A,
                        measured in situ
      TPH
f
Q
                257 nm     266 nm     278 nm      289 nm     299 nm      315 nm      327 nm     342 nm
    Cone (jig/g)
Summed fluorescence (relative units)
   Figure 13. A comparison of laboratory core sample analysis (32CPT-3) with in situ CPT-LIF
   measurements at each excitation wavelength for CPT push #6.

-------
   Lab Analysis of

    Core samples
                        LIF signal at each excitation
                             measured in situ
      TPH
s
JS
                   257 nm
266 nm
278 nm
289 nm    299 nm
315 nm
327 nm
342 nm
                                                   I
   Cone 
-------
 Lab Analysis of
  Core samples
                     LIF signal at each excitation A,
                          measured in situ
      TPH
s
JS
                      257 nm
266 nm
278 nm
289 nm    299 nm
315 nm
327 nm
342 nm
   Cone 
-------
 Lab Analysis of
  Core samples
                     LIF signal at each excitation A,
                          measured in situ
      TPH
s
JS
                      257 nm
266 nm
278 nm
289 nm    299 nm
315 nm
327 nm    342 nm
   Cone 
-------
                     299 nm
315 nm
327 nm
                     257 nm

16 -
12 -
10 -
6 -
4.


1


f
1
J
0 E+00 7 E-01 1 E+00 2 E+00
 266 nm
                                                                342 nm
278 nm
                 0 E+00  7 E-01  1 E+00 2 E+00    0 E+00 2 E-06 3 E-06 5 E-06    0 E+00 7 E-01 1 E+00  2 E+00    0 E+00 1 E-06  2 E-06  3 E-Of
289 nm

16 •
12 •
10 •
6 •
[
1
1
/
(
I
X
f
0 E+00 1 E-06 2 E-06 3 E-06
                                                                                           =
                          Summed fluorescence (relative units)
Figure 1. Depth vs. summed fluorescence for each excitation wavelength for push 1.

-------
                    299 nm
315 nm
327 nm

16 •
12 •
10 •
6 •
0 E
r"
^
>
00 5 E-07 1 E-06 2 E-06
                    257 nm
                0 E+OO 7 E-01 1 E+00 2 E+00
                                                              342 nm
 266 nm
                                i>
278 nm

,
16-

10 -
1
1
i
6 -
H
H
r


0 E+00 3 E-08 6 E-08 9 E-08
289 nm
                                                           0 E+00  1 E-06  2 E-06  3 E-06
                         Summed fluorescence (relative units)
Figure 2. Depth vs. summed fluorescence for each excitation wavelength for push 2.

-------
                     299 nm
315 nm
327 nm
                                                                 342 nm
                     257 nm
             -C
             +J

              Q<
              4>

             Q

16
12
10
6



)
0 E+00 7 E-01 1 E+00 2 E+00
                                 0 E+00  2 E-07 4 E
 266 nm
278 nm
                 0 E+00 7 E-01 1 E+00  2 E+00     0 E+00  2 E-06  3 E-06  5 E-06     0 E+00  7 E-01 1 E+00  2 E+00    0 E+00 1 E-06  2 E-06  3 E-Of
18 •
16 •
12 •
10 •
6 •
0 •
0 E+



\
00 7 E-01 1 E+00 2 E+00
289 nm


12 •
10 •
6 •
I
I


i
0 E+00 1 E-06 2 E-06 3 E-06
                          Summed fluorescence (relative units)


Figure 3. Depth vs. summed fluorescence for each excitation wavelength for push 3.

-------
                   299 nm
315 nm
327 nm
                                                          342 nm










£H
-C
•^^

Q



18 •
16 •
14 •
10 •
8 •
6 •
4
2 •
0 E+

f


\
19


1








00 1 E-07 2 E-07 3 E-07
257 nm
20
16
14
12
10





;
8I
6
0
0
|

E+00 7 E-01 1 E+00 2 E+00
















20 -j.
18 •
16 •
14 •
10 •
8 •
6 •
4-
2 •
0 •
0 E+



>




!
00 3 E-08 6 E-08 9 E-08








266 nm
20 -
16 -
14 -
12 -
10 -
8-
6 -
0 -
0 E






1
'
1
00 2 E-06 3 E-06 5 E-06
























18 -
16 -
14-
10 -
8-
6 -
4 -

0 E+





V

'
00 6 E-08 1 E-07 2 E-07








278 nm
20 •
18 •
16 •
14 •
12 •
10 •
8 •
6 •
2 •
0 •
0 E








,
00 7 E-01 1 E+00 2 E+







00








1

8 -
16 -
14 -
10 -
8:
6 -
4-

2 A







'»








0 E+00 2 E-08 4 E-08 6 E-08








289 nm
20
18
16
14
12
10
8
6
2
0
0






L
\
1
+00 1 E-06 2 E-06 3 E-06
                       Summed fluorescence (relative units)
Figure 4. Depth vs. summed fluorescence for each excitation wavelength for push 4.

-------
                      299 nm
315 nm
327 nm
                   0 E+00  1 E-06  2 E-06  3 E-06
                      257 nm
              -C
              +J

              Q<
              4>

              Q
                  -3 E-01 2 E-01 7 E-01  1 E+00
                                  0 E+00  1 E-07  2 E-07  4 E-l
 266 nm
                                                                   342 nm
278 nm
                                   0 E+00  2 E-06  3 E-06  5 E-06     0 E+00 7 E-01  1 E+00  2 E+00    0 E+00  1 E-06 2 E-06  3 E-06
289 nm
                           Summed fluorescence (relative units)


Figure 5. Depth vs. summed fluorescence for each excitation wavelength for push 5.

-------
                     299 nm
                    257 nm
                0 E+OO  2 E+OO  4 E+OO  6 E+0:
315 nm
 266 nm
327 nm
278 nm
                                               i
                                               i
                                                               342 nm
                                                            0 E+OO 3 E-OS  6 E-08  9 E-OS
289 nm
                                                             0 E+OO  1 E-06  2 E-06 3 E-0£
                         Summed fluorescence (relative units)
Figure 6. Depth vs. summed fluorescence for each excitation wavelength for push 6.

-------
                    299 nm
315 nm
327 nm
                                                             342 nm
                 0 E+00 3 E-07 6 E-07 9 E-07
                    257 nm
 266 nm
                               0 E+00 2 E-06  3 E-06  5 E-06

18 •
16 •
14 •
12 •
10 •
8 •
6 •
2 •
0 E
k
f
I
\
\
>



00 1 E-08 2 E-08 3 E-08
              289 nm
                         Summed fluorescence (relative units)
Figure 7. Depth vs. summed fluorescence for each excitation wavelength for push 7.

-------
                    299 nm
315 nm
327 nm
                 0 E+00 3 E-07  6 E-07 9 E-07
                   257 nm
 266 nm

18 •
16 •
14 •
12 •
10 •
6 •
2 •

-

H
H
1
H
i
i,
0 E+00 2 E-06 3 E-06 5 E-06
                                                            342 nm

18 •
16 •
14 •
1
12!
10 !
8 •

6 •
1
1
2 !

1
-i-,
h
L
1
i
\
i
H
-\
-\
4-t
0 E+00 1 E-08 2 E-08 3 E-08
              289 nm
                        Summed fluorescence (relative units)
Figure 8. Depth vs. summed fluorescence for each excitation wavelength for push 8.

-------
                     299 nm
             -C
             +J

             Q<
             4>

             Q
                  0 E+00 3 E-07 5 E-07  8 E-07
                    257 nm
315 nm
 266 nm
                                0 E+OO 1 E-06  2 E-06  3 E-06
327 nm
                                                               342 nm
                                                             0 E+00 3 E-09 6 E-09  9 E-09
               289 nm
                         Summed fluorescence (relative units)


Figure 9. Depth vs. summed fluorescence for each excitation wavelength for push 9.

-------
                   299 nm
315 nm
327 nm
                                                           342 nm
                   257 nm
 266 nm

14 •
12 •
10 •
8 •


2 • H
H
H
H
H
V
I ,
— <— 1
— — 1
	 1
— ' — 1
0 E+00 1 E-06 2 E-06 3 E-06
              289 nm
                                                          r
                        Summed fluorescence (relative units)
Figure 10. Depth vs. summed fluorescence for each excitation wavelength for push 10.

-------
                    299 nm
315 nm
327 nm
                  0 E+00  3 E-07  5 E-0"
                    257 nm
 266 nm
                                0 E+00  1 E-06  2 E-06 3 E-06
                                               4 • |-Js-

                                               I
                                                               342 nm
                                                            0 E+00 1 E-08 2 E-OS 4 E-l
               289 nm
                         Summed fluorescence (relative units)
Figure 11. Depth vs. summed fluorescence for each excitation wavelength for push 11,

-------
                    299 nm
315 nm
327 nm
                  0 E+OO  3 E-07  5 E-07  8 E-IX
                    257 nm
 266 nm
                                0 E+OO  2 E-06 4 E-06 6 E-06
                                                               342 nm
                                                            0 E+OO  1 E-OS  2 E-08  3 E-Oi
               289 nm
                         Summed fluorescence (relative units)
Figure 12. Depth vs. summed fluorescence for each excitation wavelength for push 12.

-------
                     299 nm
                  0 E+OO  3 E-07  5 E-07  8 E-IX
                    257 nm
             -C
             +J

             Q<
             4>

             Q
315 nm
 266 nm
                                0 E+OO 2 E-06 4 E-06 6 E-06
327 nm
                                                               342 nm
                                                             0 E+OO 1 E-OS 2 E-08  3 E-Oi
               289 nm
                         Summed fluorescence (relative units)


Figure 13. Depth vs. summed fluorescence for each excitation wavelength for push 13.

-------
                     299 nm
315 nm
327 nm
                                                                342 nm
                  0 E+00  1 E-07  2 E-07  3 E-07
                     257 nm
                                 0 E+00 3 E-08 5 E-08 8 E-08
                                                              0 E+00 1 E-08  2 E-08  3 E-08
 266 nm
               289 nm
             -C
             +J

             Q<
             4>

             Q
                                 E+00  2 E-06 4 E-06 6 E-06
                          Summed fluorescence (relative units)


Figure 14. Depth vs. summed fluorescence for each excitation wavelength for push 14.

-------
                    299 nm
                  0 E+00  1 E-07  2 E-07 3 E-07
                    257 nm
             €H

             -C

315 nm
 266 nm
                                0 E+00   2 E-06  4 E-06
327 nm
                                                               342 nm
                                                            0 E+00  1 E-08  2 E-08  3 E-Oi
               289 nm
                         Summed fluorescence (relative units)
Figure 15. Depth vs. summed fluorescence for each excitation wavelength for push 16.

-------
                    299 nm
                 0 E+OO  2 E-07  3 E-07 5 E-07
                   257 nm
315 nm
 266 nm
327 nm

14 •
12 •
10 •
6 •
2 •
i



1 '
I
0 E+OO 1 E-06 2 E-06 3 E-06
                                                            342 nm
              289 nm
                        Summed fluorescence (relative units)
Figure 16. Depth vs. summed fluorescence for each excitation wavelength for push 17.

-------
                   299 nm
315 nm
327 nm
                                                           342 nm
                   257 nm
                 r
 266 nm
              289 nm
                        Summed fluorescence (relative units)
Figure 17. Depth vs. summed fluorescence for each excitation wavelength for push 18.

-------
                    299 nm
                 0 E+OO 7 E-08  1 E-07  2 E-07
                    257 nm
315 nm
 266 nm
327 nm
                                                             342 nm
                                                           0 E+OO 3 E-09  6 E-09  9 E-09
              289 nm
                         Summed fluorescence (relative units)
Figure 18. Depth vs. summed fluorescence for each excitation wavelength for push 19.

-------
                    299 nm
                  0 E+00  2 E-07  4 E-07 6 E-07
                    257 nm
315 nm
 266 nm
                                0 E+OO  2 E-06  3 E-06 5 E-06
327 nm
                                                               342 nm
                                                            0 E+00 6 E-09 1 E-OS 2 E-08
               289 nm
                         Summed fluorescence (relative units)
Figure 19. Depth vs. summed fluorescence for each excitation wavelength for push 20.

-------
                    299 nm

16 •
12 •
10 •
6 •



k
0 E+00 2 E-07 4 E-07 6 E-07
                    257 nm
315 nm
327 nm
                               0 E+00 3 E-08 5 E-08 8 E-08
 266 nm
                               0 E+00 2 E-06  3 E-06  5 E-06
                                                             342 nm
              289 nm
                         Summed fluorescence (relative units)
Figure 20. Depth vs. summed fluorescence for each excitation wavelength for push 21,

-------