September 2006
                            NSF 06/ARS1/EPADWCTR
                                EPA/600/R-06/099
Environmental Technology
Verification Report
Removal of Arsenic in Drinking Water

ARS USA,  LLC
ARS CFU-50 ARC Electroflocculation and
Filtration Water Treatment System
                Prepared by
             NSF International
          Under a Cooperative Agreement with
          U.S. Environmental Protection Agency

-------
         THE ENVIRONMENTAL TECHNOLOGY VERIFICATION
                                      PROGRAM
    U.S. Environmental Protection Agency                                                NSF International

                      ETV Joint Verification  Statement
    TECHNOLOGY TYPE:   ELECTROFLOCCULATION AND MEDIA FILTRATION
                            USED IN DRINKING WATER TREATMENT SYSTEMS
    APPLICATION:         REMOVAL OF ARSENIC IN DRINKING WATER
    TECHNOLOGY NAME:  ARS CFU-50 APC ELECTROFLOCCULATION AND
                            FILTRATION WATER TREATMENT SYSTEM

    COMPANY:             ARS USA, LLC
    ADDRESS:              PO Box 1170                       PHONE:  (505)771-4344
                            Bernalillo, NM 87004              FAX:     (505) 771-4345
    WEB SITE:              www.arsusa.com
    EMAIL:                 info@arsusa.com
The U.S. Environmental Protection Agency (EPA) supports the Environmental Technology Verification
(ETV)  Program to facilitate the deployment of innovative  or improved environmental technologies
through performance verification and dissemination of information. The goal of the ETV Program is to
further environmental protection by accelerating  the acceptance and use of improved and more cost-
effective technologies.  ETV seeks to achieve this goal by providing high-quality, peer-reviewed data on
technology performance to those involved in the design, distribution, permitting, purchase, and  use of
environmental technologies.
ETV works  in partnership with recognized standards  and testing organizations, stakeholder groups
(consisting of buyers, vendor organizations, and permitters), and with the full participation of individual
technology developers.  The program evaluates the performance of innovative technologies by developing
test plans that are  responsive to the needs  of  stakeholders,  conducting field or laboratory  tests (as
appropriate), collecting and analyzing data, and preparing peer-reviewed reports.  All evaluations are
conducted in accordance with rigorous quality assurance protocols to  ensure that data of known and
adequate quality are generated and that the results are defensible.
NSF International  (NSF), in cooperation  with the EPA, operates the Drinking Water Systems (DWS)
Center, one of six technology areas under the ETV  Program.  The DWS  Center recently evaluated the
performance of an electroflocculation and media filtration system for the removal of arsenic from
drinking water. This verification statement provides  a summary of the test results for the ARS CFU-50
APC Electroflocculation and Filtration Water Treatment System (ARS CFU-50 APC). The NSF Drinking
Water Treatment Systems Laboratory  (DWTS) was the field testing organization (FTO) that performed
the verification testing.  The verification report contains a comprehensive description of the complete
verification test.
06/ARS1/EPADWCTR    The accompanying notice is an integral part of this verification statement.        September 2006

                                        VS-i

-------
ABSTRACT
Verification testing of the ARS CFU-50 APC Electroflocculation and Filtration Water Treatment System
(ARS CFU-50 APC) for arsenic removal was conducted at the Town of Bernalillo Well #3 site from April
18 through May 2, 2006.  The source water was chlorinated groundwater from two supply wells, and the
feed water for the verification test was withdrawn from the pressure tank at the site. Verification testing
was  conducted at the operating conditions specified by the manufacturer.  The feed water, with a pH in
the range of 7.6 to 7.9, was pumped into a reaction vessel where electricity is applied to aluminum and
graphite plates to create flocculent to which arsenic adsorbs.  When operated under the manufacturer's
specified conditions at this site, at  an average flow rate of 32.1 gallons per minute  (gpm), the ARS
CFU-50 APC reduced the total arsenic concentration from an average of 12 micrograms per liter ((ig/L)
in the feed (untreated) water to 6 (ig/L in the filtrate (treated) water.
TECHNOLOGY DESCRIPTION
The following technology description was provided by the manufacturer and has not been verified.
The  ARS  CFU-50 APC is a standard, full-scale,  modular system for the removal of arsenic and other
contaminants from water. The ARS CFU-50 APC is a self-contained, complete system that connects to a
water supply source. If the source is  not pressurized, a pump, supplied with the unit, is used to pump the
water through the treatment system.  The ARS CFU-50  APC requires a three-phase 480-volt AC electric
power source to operate  the reaction vessel,  programmable logic  controller (PLC), and  ancillary
equipment. The  system used for this test is  designed to treat flows up to a maximum flow rate  of
approximately 35 gpm (50,000 gallons per day [gpd]),  from either a pressurized or unpressurized water
source.
Untreated/contaminated water enters the unit through a regulated influent pipe. The flocculent generation
and decontamination process occurs in the reaction vessel in a continuous process. Flocculent particles in
the holding pipe/tank are subject to further growth and reaction after the electrolytic process. Sand filters
separate the flocculent from the treated water. The filter surfaces are cleaned by automatic backwashing,
and the flocculation sludge is flushed into the floe water reservoir tank. The  low volume, thickened
flocculation sludge accumulated in the floe water reservoir tank is pumped into the filter press by a pump,
where it is pressed into a filter cake. After the treated water passes through the filter press, it is stored in
the clean water tank for later use in filter backwashing and rinsing. As the clean water tank level reaches
its maximum level, it is pumped out of the unit through the filtrate water pipe.
The ARS CFU-50 APC treatment system is fully automated and programmed to control all aspects of the
treatment  and filter operation. The control system automatically initiates backwash cycles based on  an
inlet pressure level set by the operator.  The backwash  cycle time is a  fixed time duration that is
programmed in the PLC. The control system monitors data from the system operation. This information is
available to the on-site operator.
VERIFICATION TESTING DESCRIPTION
Test Site
The  Bernalillo Well #3 site is a fenced property that includes a building that houses the well pump and
chlorination equipment, a primary storage  tank (approximately 1,000,000 gallons [gal]), and a secondary
storage tank (approximately 200,000 gal).  Water pumped at the site is a mixture from two wells, both of
which pump water from the Rio Grande Group aquifer.  The  average daily water use for the Town of
Bernalillo  is approximately 2,000,000 gpd. Water quality data based on data collected between June 2002
and March 2004  shows total arsenic in the combined well water ranges from  14 to  68  (ig/L and the
primary arsenic  species is  arsenic (V).   The water has a total hardness  of approximately 70 to  90
milligrams per liter (mg/L) as CaCO3 and the pH is approximately 7.3.
06/ARS1/EPADWCTR    The accompanying notice is an integral part of this verification statement.         September 2006

                                         VS-ii

-------
Methods and Procedures

Operations, sampling, and analyses were performed in accordance with the Product Specific Test Plan
(PSTP) developed and approved for this verification test. The PSTP included a Quality Assurance Project
Plan (QAPP) to assure the quality of the data collected  and to provide an accurate evaluation of the
treatment system under field conditions. Testing included characterization of the feed water, an arsenic
loss test (no electricity supplied to the reaction vessel), and a 14-day verification test.
The verification test was performed from April 18, through May 1, 2006. The ARS CFU-50 APC was
operated for the 14-day verification test by using water supplied from the Town of Bernalillo. Flow rate,
production volume, water temperature, and system pressure were monitored and recorded daily. Feed and
filtrate (treated) water samples were analyzed on-site for pH, temperature, turbidity, free and total residual
chlorine, color, and dissolved oxygen (DO) by the field operator.  Grab samples were collected and
delivered to the NSF  Analytical  Laboratory  and were  analyzed  for alkalinity,  aluminum,  calcium,
magnesium, iron, manganese, sulfate, chloride, total organic carbon (TOC), total suspended solids (TSS),
and fluoride.  Samples for total  arsenic  were collected daily, plus 14  samples were collected during a
48-hour intensive survey.  In addition to the samples for total arsenic, arsenic samples were speciated
during the test to determine the soluble arsenic concentration and the concentrations of arsenic (III) and
the arsenic (V) present in the soluble fraction.
Complete descriptions of the verification testing results and quality assurance/quality control (QA/QC)
procedures are included in the verification report.
VERIFICATION OF PERFORMANCE
System Operation

ARS performed the system startup  and shakedown testing, which included optimization of the electrical
feed rates (30 amps) to the reaction vessel. The verification test was conducted under the manufacturer's
specified operating conditions. The backwash system was  set to backwash when the pressure differential
across the filter exceeded 15 pounds per square inch (psi).
System pressure was monitored at the filter influent and  filtrate.  Head loss fluctuated between 6.4 and
15.9 psi  during the  inspections.  The ARS  CFU-50 APC PLC was not programmed to  record pressure
differentials at the start  of backwash cycles, so the pressure differential  evaluation for this verification
was limited to whether the differential exceeded 15 psi during the time the FTO personnel inspected the
device.
During the test, there were a total of four incidences (April 20, 21, 28,  and 30) where a sensor triggered
the PLC to shut down operations.  During each incident,  the sensor indicated that either the floe water
reservoir tank had exceeded capacity  or the filter press alarm went off.  In each instance, the filter press
had clogged to a point where it  was prohibiting sufficient filtration to  maintain  the device's rated
throughput.  ARS personnel  recommended that the filter press be cleaned a minimum of once every  24
hours to prevent the ARS CFU-50 APC from automatically shutting down. After each shutdown incident,
FTO personnel cleaned the filter press and resumed operation in accordance with the startup procedures
outlined in the ARS Operations and Maintenance (O&M) manual.  As a  result of these incidents, the ARS
CFU-50 APC experienced approximately 36 hours of downtime during the 14-day verification test.
The filtrate flow rate was  32.1 gpm over the 14 days. The total filtrate volume produced each day was
also consistent, except for those days when operating time was lost due to the filter press alarm shutting
down the system.
Water Quality Results
The results of total  arsenic analyses are  shown in Figure VS-1.  The feed water total arsenic averaged
12 (ig/L with most of the arsenic as arsenic (III), but with some arsenic (V) also present. The filtrate water
total arsenic concentration averaged 6 (ig/L. The data collected during the 48-hour intensive survey were
06/ARS1/EPADWCTR   The accompanying notice is an integral part of this verification statement.         September 2006

                                          VS-iii

-------
consistent with the data collected each day during the verification test. There was no indication of any
transient or short time changes in the arsenic concentration or in any other monitored parameters.
      16 -,
                        *%   -%
%
%
                                             Date
                                         -Feed —•—Filtrate
Figure VS-1. Total Arsenic Results.

The feed and filtrate water alkalinity averaged 130 mg/L as CaCO3, indicating that the treatment process
had no  impact  on the  alkalinity.   The pH of the feed and filtrate water had a median value of 7.7.
Aluminum was detected in four of the 14 feed water samples, at concentrations ranging from 13 to
84 (ig/L, while  the remaining ten feed water samples had aluminum concentrations below the 10 (ig/L
detection limit.  In the filtrate, the average aluminum concentration was 560 (ig/L, and ranged from 200 to
890 (ig/L. The average  filtrate aluminum concentration was 20 times greater than the feed water average
concentration and significantly higher than the National Secondary Drinking Water Regulation range of
50 to 200 (ig/L. Furthermore, operation of the ARS CFU-50 APC increased the turbidity levels in the
filtrate water. The feed water turbidity averaged 0.30 Nephelometric Turbidity Units (NTU), and ranged
from 0.20 to 0.45 NTU, while the filtrate water averaged 0.80 NTU, and ranged from 0.35 to 1.2 NTU.
Turbidity and aluminum data during the 48-hour intensive survey were similar to those during the 14-day
test.   The turbidity and aluminum data indicated that filtration mechanisms more efficient than those
currently utilized  in the ARS CFU-50 APC were required to bring these parameters closer to the  feed
water concentrations or within the  EPA regulations. The ARS CFU-50 APC had little or no impact on
free chlorine, total chlorine, DO, chloride, sulfate, TOC, fluoride, calcium, or magnesium concentrations.
Manganese and iron concentrations were consistently below detection  limits in both the feed and filtrate
water.
Backwash was  initiated automatically based on pressure differential.   Backwash waste was treated by a
filter press designed to remove the solids  (floe) from the backwash water.  The filtrate from the filter
press was transferred back to the reaction vessel for re-treatment.  The backwash cycle was set for a fixed
time duration of 120 seconds for backwash and  30  seconds for rinsing.  The combined backwash and
rinsing resulted in approximately 250 gallons of waste per backwash sequence. Solids retained in the
filter press were removed manually during filter press maintenance. At the end of testing, approximately
572,550  gallons of water were treated, and approximately 1,425  pounds of solids  (wetted floe)  was
created.  This calculates to an approximate suspended solids concentration of 300 mg/L.  The backwash
06/ARS1/EPADWCTR    The accompanying notice is an integral part of this verification statement.

                                          VS-iv
                                          September 2006

-------
solids were  not considered a hazardous waste, based on Toxicity Characteristic Leaching Procedure
(TCLP) metals  analyses, which were below the regulatory limits under the Resource Conservation and
Recovery Act (RCRA).
Operation and Maintenance Results

The ARS CFU-50 APC was found to be easy to operate and required little time for daily maintenance.
The field staff was on-site for two to three hours per day. Most of the time on-site was spent performing
field activities, including flow checks, calibrations, cleaning the filter press, and other verification-related
activities.
The ARS CFU-50 APC O&M manual provides a  detailed description of the system, appropriate safety
precautions,  and detailed descriptions of operating procedures, capability  and operation of the computer
control system, and specific instructions for utility operators. The maintenance section of the  manual
includes  some descriptions  of required maintenance, but refers the reader to the individual equipment
literature supplied by  the various pump and instrument manufacturers. A review of the O&M  manual
shows that the manual is well organized and easy to read.
The ARS CFU-50 APC was equipped with two sand filters, so that one filter could be in operation while
the other was  in backwash mode or  standby.  During the testing at this installation,  there were no
conditions where the pressure  differential across both sand filters required  that both filters backwashed at
the same time.  Issues regarding the efficacy of the filtration process, as shown in the  aluminum and
turbidity data, were noted during the verification test.
Backwash waste was treated by a filter press designed to remove the solids from the backwash water.
During the testing, when the flocculent caked in the filter press to a point where water would no longer
pass through it, the PLC shut down the entire system, as it was programmed to do. When this occurred,
field personnel  cleaned the filter  press and restarted the system. Verification testing substantiated the
importance of the filter press  and its appropriate maintenance as a critical aspect of the function of the
ARS CFU-50 APC.
The system PLC was designed to operate and monitor many of the operating functions of the device.  The
PLC readings were easy to use, but required an understanding of the PLC operating  keys to display the
readings. The PLC was not programmed to record data, so readouts on component performance,  such as
flow, pressure, and electrical settings had to be monitored and recorded manually. Because the PLC did
not record data, information regarding the duration of filter runs, frequency of backwash cycles,  and the
pressure differentials across the sand filters could not be accurately recorded. The PLC was designed to
shut the entire system  down in the event any sensor recorded a condition outside preset operating limits.
This condition was experienced four times during the verification.  The cause of each shutdown was the
filter press clogging to a point where water could not pass through it at the system's rated throughput.
During each shutdown condition, after the filter press was cleaned, the alarm conditions in the PLC were
cleared and the system was restarted without difficulty.
Electrical power consumption was estimated based  on the floe pump, clean water pump, backwash pump,
reaction vessel, waste pump,  and miscellaneous other devices (air compressor, PLC, lights, etc.). The
power consumption was estimated to be 4.2 kilowatt hours (KwH).
Quality Assurance/Quality Control
NSF provided technical and  QA  oversight of the verification testing  as described in the verification
report, including an audit of nearly 100% of the data. The NSF QA department conducted a technical
systems audit during testing to ensure the testing was in compliance with the test plan and performed a
QA review of the analytical data.  A complete description of the QA/QC procedures is provided in the
verification report.
06/ARS1/EPADWCTR    The accompanying notice is an integral part of this verification statement.        September 2006

                                          VS-v

-------
   Original signed by                                    Original signed by
   Sally Gutierrez	September 22, 2006         Robert Ferguson   September 12, 2006
    Sally Gutierrez                     Date            Robert Ferguson             Date
    Director                                           Vice President
    National Risk Management Research Laboratory       Water Systems
    Office of Research and Development                 NSF International
    United States Environmental Protection Agency
    NOTICE:  Verifications are based on an evaluation of technology performance under specific,
    predetermined criteria and the appropriate quality assurance procedures.  EPA and NSF make no
    expressed or implied warranties as to the performance of the technology and do not certify that a
    technology will always operate as verified.  The end-user is solely responsible for complying with
    any and all applicable federal, state, and local requirements.  Mention of corporate names, trade
    names, or commercial products does not constitute endorsement or recommendation for use of
    specific products.  This report is not an NSF Certification  of the specific product mentioned
    herein.
       Availability of Supporting Documents
       Copies of the ETV Protocol for Equipment Verification  Testing for Arsenic Removal
       dated September 2003, the product-specific test plan, the verification statement, and the
       verification  report  (NSF  Report  #06/ARS1/EPADWCTR)  are  available from  the
       following sources:
       (NOTE: Appendices are not included in the verification report.  Appendices are available
       from NSF upon request.)

       1.   ETV Drinking Water Systems Center Manager (order hard copy)
           NSF International
           P.O.  Box 130140
           Ann Arbor, Michigan 48113-0140
       2.   NSF web site: http://www.nsf.org/info/etv (electronic copy)
       3.   EPA web site: http://www.epa.gov/etv (electronic copy)
06/ARS1/EPADWCTR    The accompanying notice is an integral part of this verification statement.         September 2006

                                         VS-vi

-------
                                                         September 2006
        Environmental Technology Verification Report
            Removal of Arsenic in Drinking Water

                        ARS USA, LLC
      ARS CFU-50 APC Electroflocculation and Filtration
                   Water Treatment System
                          Prepared for:

                        NSF International
                    Ann Arbor, Michigan 48105
                           Prepared by:

                        NSF International
Under a cooperative agreement with the U.S. Environmental Protection Agency
                  Jeffrey Q. Adams, Project Officer
            National Risk Management Research Laboratory
                U.S. Environmental Protection Agency
                      Cincinnati, Ohio 45268

-------
                                       Notice

The  U.S.  Environmental Protection  Agency  (EPA),  through its  Office  of  Research and
Development, has financially supported and collaborated with NSF International (NSF) under
Cooperative Agreement No. R-82833301. This verification effort was supported by the Drinking
Water  Systems  (DWS) Center, operating  under the Environmental Technology Verification
(ETV)  Program. This document has  been peer reviewed,  reviewed by NSF and EPA, and
recommended for public release.
                                         11

-------
                                       Foreword

The U.S. Environmental Protection Agency (EPA) is charged by Congress with protecting the
Nation's land, air, and water resources. Under a mandate of national  environmental laws, the
Agency strives to formulate and implement actions leading to a compatible balance between
human activities  and the ability of natural systems to support and nurture life.  To meet this
mandate,  EPA's  research  program  is providing  data  and  technical  support  for  solving
environmental problems today and building a science knowledge base necessary to manage our
ecological resources wisely, understand how pollutants affect our health, and prevent or reduce
environmental risks in the future.

The National  Risk Management Research Laboratory (NRMRL)  is the Agency's center for
investigation of technological  and  management approaches for preventing and reducing risks
from pollution that threaten human health  and the environment. The focus of the Laboratory's
research program is on methods and  their cost-effectiveness for prevention  and control  of
pollution to air, land, water, and subsurface resources; protection of water quality in public water
systems; remediation of contaminated sites, sediments and ground water; prevention and control
of indoor air pollution; and restoration of  ecosystems.  NRMRL collaborates with both public
and private  sector partners to foster technologies  that reduce the  cost  of compliance and to
anticipate emerging problems. NRMRL's research provides solutions to environmental problems
by: developing and promoting technologies that protect and improve the environment; advancing
scientific and engineering information to support regulatory and policy  decisions; and providing
the technical  support  and  information transfer  to ensure implementation of environmental
regulations and strategies at the national, state, and community levels.

This publication has been produced as part of the Laboratory's strategic long-term research plan.
It is published and made available by EPA's Office of Research and Development to assist the
user community and to link researchers with their clients.
                                         Sally Gutierrez, Director
                                         National Risk Management Research Laboratory
                                           in

-------
                                   Table of Contents

Verification Statement	VS-i
Notice	ii
Foreword	iii
Table of Contents	iv
List of Figures	v
List of Tables	v
Appendices	vi
Abbreviations and Acronyms	vii
Acknowledgements	viii
Chapter 1 Introduction	1
   1.1    ETV Purpose and Program Operation	1
   1.2    Testing Participants and Responsibilities	1
     1.2.1    NSF International	2
     1.2.2    Field Testing Organization	2
     1.2.3    Manufacturer	3
     1.2.4    Analytical Laboratory	3
     1.2.5    U.S. Environmental Protection Agency	4
   1.3    Verification Testing Site	4
     1.3.1     Site Background Information	4
     1.3.2     Source/Feed Water Quality	4
     1.3.3    Test Site Description	5
Chapter 2 Equipment Capabilities and Description	7
   2.1    Description of Equipment	7
   2.2    Engineering and  Scientific  Concepts	7
     2.2.1    Physicochemical Efficient Mechanisms	8
   2.3    Description of Treatment Train and Unit Processes	8
   2.4    Description of Physical Construction and Components	12
   2.5    Chemical Consumption and Production of Waste Material	12
     2.5.1    Chemical Consumption	12
     2.5.2    Waste Production and Physical and Chemical Nature of Wastes	13
   2.6    Licensing Requirements	13
   2.7    Statement of Performance Objectives	13
   2.8    Advantages of the ARS CFU-50 APC  Process	14
   2.9    Potential Limitations of the Equipment	14
Chapters Methods and Procedures	15
   3.1    Quantitative and  Qualitative Evaluation Criteria	15
   3.2    Key Water Quality Parameters	15
   3.3    Operations and Maintenance	15
   3.4    Environmental Technology Verification Testing Plan	16
     3.4.1    Task A:  Raw Water Characterization	16
     3.4.2    TaskB:  Arsenic Loss Test	16
     3.4.3    Task C:  Verification Test Procedures	16
   3.5    Operation and Maintenance	18
     3.5.1    Operability Evaluation	18
                                           iv

-------
Chapter 4 Results and Discussion	19
  4.1    Introduction	19
  4.2    Equipment Installation, Start-up, and Shakedown	19
     4.2.1    Flow Measurement	19
  4.3    Task A: Raw Water Characterization	20
  4.4    Task B: Initial Test Runs	21
     4.4.1    Arsenic Loss Test	21
  4.5    TaskC: Verification Test	24
     4.5.1    Operating Results	24
     4.5.2    Arsenic Results	25
     4.5.3    Feed and  Filtrate Water Quality Results	29
  4.6    Operations and Maintenance Findings	40
     4.6.1    Electrical Consumption	41
     4.6.2    Sand Filters	41
     4.6.3    Filter Press	42
     4.6.4    Backwash Water Frequency and Quality	42
     4.6.5    Programmable Logic Controller	43
  4.7    Quality Assurance/Quality Control	44
     4.7.1    Documentation	44
     4.7.2    Quality Audits	44
     4.7.3    Data Quality Indicators	45
     4.7.4    Effect of  Sample Preservative on Arsenic Speciation	52
     4.7.5    Deviations from PSTP	53
Chapters References	54

                                     List of Figures

Figure 2-1. ARS CFU-50 APC schematic view	9
Figure 2-2. ARS CFU-50 APC right isometric view	11
Figure 2-3. ARS CFU-50 APC skid-mounted unit photograph	11
Figure 4-1. Verification test daily arsenic results	29
Figure 4-2. Verification test pH results	31
Figure 4-3. Verification test turbidity results	33
Figure 4-4. Verification test alkalinity results	35
Figure 4-5. Verification test aluminum results	37

                                      List of Tables

Table 1-1. Raw Water Quality Data	5
Table 2-1. Test System Operating Conditions	10
Table 2-2. ARS  CFU-50 APC System Specifications	10
Table 3-1. Key Filtrate Water Quality Parameters	15
Table 4-1. Feed  Water Characterization Data - April 18, 2006	20
Table 4-2. TaskB Arsenic Loss Test Operating Data	21
Table 4-3. Task  B Arsenic Loss Test Water Quality Results	23
Table 4-4. Operating Data	25

-------
Table 4
Table 4
Table 4
Table 4
Table 4
Table 4
Table 4
Table 4
Table 4
Table 4
Table 4
Table 4
Table 4
Table 4
Table 4
Table 4
Table 4
Table 4
Table 4
Table 4
Table 4
Table 4
-5. Daily Total Arsenic Results (|ig/L)	26
-6. Total Arsenic Results for 48-Hour Intensive Survey (|ig/L)	27
•7. Arsenic Speciation Data (|ig/L)	28
•8. pH Results (S.U.)	30
•9. pH Results for the 48-Hour Intensive Survey (S.U.)	30
•10. Bench Top Turbidity Results	32
•11. Bench Top Turbidity Results for the 48-Hour Intensive Survey	33
•12. Alkalinity Results	34
•13. Alkalinity Results for the 48-Hour Intensive Survey	35
•14. Aluminum Results	36
•15. Aluminum Results for the 48-Hour Intensive Survey	37
•16. Total and Free Residual Chlorine and DO	38
•17. Total and Free Residual Chlorine and DO Results for 48-Hour Survey	39
•18. Other Water Quality Parameters	40
•19. Power Consumption	41
•20. Backwash Solids - TCLP and CAWET Analyses	43
•21. Field Instrument Calibration Schedule	46
•22. Flow Meter Calibration Data	47
•23. Precision Data - Field Duplicates for Laboratory Parameters	49
•24. Precision Data - Field Duplicates for Field Parameters	50
•25. Completeness Results	52
•26. Deviations from PSTP	53
                                     Appendices

Appendix A - Operation and Maintenance Manual
Appendix B - Product Specific Test Plan
Appendix C - Spreadsheets
Appendix D - Field Data Logbook and Calibration Records
Appendix E - NSF Laboratory Data Reports
Appendix F - TriMatrix Laboratories Data Report for TCLP and CAWET Analyses
                                          VI

-------
                            Abbreviations and Acronyms
ARS

CAWET
°C
c.u.
DO
DWS
DWTS
EPA
ETV
ft2
FTO
gpm
gpd
hp
KwH
LCS
mg/L
mm
MSDS
NC
NIST
NR
NRMRL
NSF
NTU
O&M
ORD
PLC
psi
PSTP
QA/QC
QAPP
RCRA
%RSD
S.U.
TCLP
TOC
TSS
VAC
ARS USA, LLC (formerly  known  as Advanced Remediation Systems
USA, LLC)
California Waste Extraction Test
Degree Celsius
Color Units
Dissolved Oxygen
Drinking Water Systems
NSF International Drinking Water Treatment Systems Laboratory
U.S. Environmental Protection Agency
Environmental Technology Verification
Square Feet or Square Foot
Field Testing Organization
Gallon(s) Per Minute
Gallon(s) Per Day
Horsepower
Kilowatt-hour
Laboratory Control  Sample
Milligram(s) per Liter
Millimeter
Material Safety Data Sheets
Not Calculated
National Institute of Standards and Technology
Not Recorded
National Risk Management Research Laboratory
NSF International
Nephelometric Turbidity Unit(s)
Operation and Maintenance
Office of Research and Development
Programmable Logic Controller
Pounds per Square Inch
Product Specific Test Plan
Quality Assurance/Quality Control
Quality Assurance Project Plan
Resource Conservation and Recovery Act
Percent Relative Standard Deviation
Standard Units
Toxicity Characteristic Leaching Procedure
Total Organic Carbon
Total Suspended Solids
Microgram(s) per Liter
Volts Alternating Current
                                         vn

-------
                                  Acknowledgements

The Field Testing Organization (FTO),  NSF International (NSF) Drinking Water Treatment
Systems Laboratory (DWTS), was responsible for all elements in the testing sequence, including
collection of samples, calibration and check of instrumentation, data collection and analysis, data
management, data interpretation, and the preparation of this report.

       NSF International Drinking Water Treatment Systems Laboratory
       789 N. Dixboro Road
       Ann Arbor, Michigan 48105
       Contact Person: Mr. Robert Herman

The laboratory selected for the analytical work for this test was:

       NSF International Chemistry Laboratory
       789 N. Dixboro Road
       Ann Arbor, Michigan 48105
       Contact Person: Dr. Kurtis Kneen

The manufacturer of the equipment was:

       ARS USA, LLC
       PO Box 1130
       Bernalillo, NM 87004
       Contact Person: Mr. Norbert Barcena

The NSF DWTS  wishes to thank the following participants:

Mr. Bruce Bartley, Mr.  Patrick Davison, and Ms. Angela Beach of the NSF Environmental
Technology  Verification  (ETV) Drinking Water  Systems (DWS) Center for their support,
guidance, and program management.

The Town of Bernalillo staff, including Mr. Bill Plata and Mr. Les Swindle, for providing access
to the test site.

ARS for  supplying the verification test unit and support services during the start-up period. Mr.
Andrew Polnicki  coordinated building and shipping of the test unit, and worked at the site to
optimize  the operation of the system and provide training to the NSF field operators. Ms. Lauren
Bull provided logistical assistance.  Their work is greatly appreciated.
                                          Vlll

-------
                                      Chapter 1
                                     Introduction

1.1    ETV Purpose and Program Operation

The U.S. Environmental Protection Agency (EPA) has created the Environmental Technology
Verification  (ETV)  Program  to  facilitate  the deployment  of innovative  or  improved
environmental technologies through performance verification and dissemination of information.
The  goal  of the  ETV Program is to further environmental protection by accelerating the
acceptance and use of improved and more cost-effective technologies. ETV seeks to achieve this
goal by providing high-quality, peer-reviewed data on technology performance to those involved
in the design, distribution, permitting, purchase, and use of environmental technologies.

ETV works in partnership with recognized standards and testing organizations; with stakeholder
groups consisting  of buyers, vendor organizations, and permitters; and with the full participation
of individual technology developers.  The  program evaluates the performance  of innovative
technologies by developing test  plans responsive to the needs of stakeholders, conducting field
demonstrations, collecting and  analyzing data,  and preparing peer-reviewed  reports.   All
evaluations are conducted in accordance with rigorous quality assurance protocols to ensure that
data of known and adequate quality are generated and that the results are defensible.

The EPA has partnered with NSF International (NSF) under the ETV Drinking Water Systems
(DWS) Center to verify the  performance of small  drinking  water systems that serve small
communities.  A goal of verification testing  is to enhance and facilitate the acceptance of small
drinking water treatment equipment by state drinking water regulatory officials and consulting
engineers,  while  reducing the  need  for testing of  equipment at each location where the
equipment's use is contemplated. NSF meets this goal by working with manufacturers and NSF-
qualified Field Testing Organizations (FTOs) to conduct verification testing under the approved
protocols.  It is important to note that verification of the equipment does not mean the equipment
is "certified"  by NSF or "accepted" by EPA.  Rather, it recognizes that the performance of the
equipment has been determined and verified by these organizations for those conditions tested by
the FTO.

The DWS Center evaluated the  performance of the ARS CFU-50 APC Electroflocculation and
Filtration Water Treatment System (ARS CFU-50 APC), manufactured and distributed by  ARS
USA, LLC, which is a granular media  filtration system used in drinking water treatment system
applications for reduction of arsenic and dissolved iron in groundwater.  This document provides
the verification test results for the ARS  CFU-50 APC.

1.2    Testing Participants and Responsibilities

The ETV testing of the ARS CFU-50 APC  was a  cooperative effort among  the following
participants:

   •   NSF
   •   NSF International Drinking Water Treatment Systems Laboratory (DWTS)

-------
   •   ARS
   •   The Town of Bernalillo, New Mexico
   •   EPA

The  following  is  a  brief description  of  all of the ETV participants  and their  roles and
responsibilities.

1.2.1   NSF International

NSF is an independent, not-for-profit testing and certification organization dedicated to public
health and safety and to the protection of the environment.  Founded in 1946 and located in Ann
Arbor,  Michigan, NSF has been instrumental in the development of consensus standards for the
protection of public health and the environment.  NSF  also provides testing and certification
services to ensure products bearing the NSF Name, Logo and/or Mark meet those standards. The
EPA partnered with NSF to verify the performance of drinking water treatment systems through
the EPA's ETV Program.

NSF provided technical oversight of the verification testing and conducted an audit of the field
analytical and data gathering  and recording procedures.   NSF  also provided review of the
Product Specific Test Plan (PSTP) as well as this report.

Contact Information:
       NSF International
       789 N. Dixboro Road
       Ann Arbor,  Michigan 48105
       Contact: Bruce Bartley, Project Manager
       Phone: (734) 769-8010
       Fax: (734) 769-0109
       Email: bartley@nsf.org

1.2.2   Field Testing Organization

The DWTS conducted the verification testing of the ARS CFU-50 APC. The DWTS is an NSF-
qualified FTO for the ETV DWS Center.

The FTO provided all needed logistical support, established a communications  network, and
scheduled and coordinated activities of all participants. The FTO was responsible for ensuring
the testing location and feed water conditions were such that the verification testing could meet
its stated objectives.   The FTO prepared the PSTP;  oversaw the pilot testing;  managed,
evaluated, interpreted, and reported on the data generated by the testing; and evaluated and
reported  on the performance of the technology.  The FTO was responsible for completing the
raw water characterization testing, monitoring the ARS CFU-50 APC during the arsenic loss
testing (24 hour test), and conducting the verification test over 14 calendar days.

DWTS employees conducted the on-site analyses and data recording during the test. The FTO's
Project Manager and Project Director provided oversight of the daily tests.

-------
Contact Information:
       NSF International Drinking Water Treatment Systems Laboratory
       789 N. Dixboro Road
       Ann Arbor, Michigan 48105
       Contact Person: Mr. Robert Herman
       Phone: (734) 769-5349
       Fax: (734) 827-7143
       Email: herman@nsf.org

1.2.3   Manufacturer

The treatment system was the  ARS CFU-50 APC  Electroflocculation  and Filtration Water
Treatment System  for the removal of arsenic from  drinking water.   The manufacturer was
responsible for supplying a field-ready electroflocculation and filtration system equipped with all
necessary  components,  including treatment  equipment,  instrumentation and controls, and an
operation and maintenance (O&M) manual.   The manufacturer was responsible for providing
logistical and technical  support, as  needed, as well as technical assistance to the FTO during
operation and monitoring of the equipment undergoing field verification testing.

Contact Information:
       ARS USA, LLC
       PO Box 1170
       Bernalillo, NM 87004
       Contact Person: Mr. Norbert Barcena
       Phone: (505) 771-4344
       Fax:(505) 771-4345
       Email: norbert@arsusa.com

1.2.4   Analytical Laboratory

The NSF Chemistry Laboratory in Ann Arbor, Michigan performed all water quality analyses.

Contact Information:
       NSF International Chemistry Laboratory
       789 N. Dixboro Road
       Ann Arbor, Michigan 48105
       Contact Person: Dr. Kurtis Kneen
       Phone: (734) 827-6874
       Fax: (734) 827-7765
       Email: kneen@nsf.org

-------
Backwash toxicity analyses were performed by:

Contact Information:
       TriMatrix Laboratories, Inc.
       5555 Glenwood Hills Parkway, SE
       Grand Rapids, Michigan 49588
       Phone:(810)220-2075
       Fax:(810)220-2803
       Contact: Mr. Michael W. Movinski, Vice President, Sales and Marketing
       Email: mmtrimatrix@comcast.net

1.2.5   U.S. Environmental Protection Agency

The EPA, through its Office of Research and Development (ORD), has financially supported and
collaborated with NSF under Cooperative Agreement No. R-82833301.  This verification effort
was supported by the DWS Center operating under the ETV Program.  This document has been
peer reviewed, reviewed by NSF and EPA,  and recommended for public release.

1.3    Verification Testing Site

1.3.1   Site Background Information

The Bernalillo Well #3 site is less than one acre and includes a two-room building which houses
the well pump in one room  and the chlorination equipment in the other room. The site also
includes a primary storage tank (approximately 1 million gallons) and secondary  storage tank
(approximately 200,000  gallons).  The two tanks are connected  to each  other as well as  to
Well #3 and Well #4. The water storage  tanks are fixed wall  tanks that do not have bladder
inserts. When the water level in the storage tank drops below a preset level, Well #3 is activated
to supplement water from Well #4. Once a preset limit is met, the pumps shut off. Chlorine is
immediately added to the water from the wells prior to delivery to the storage tank.

The average daily water use for the Town of Bernalillo is approximately two million gallons per
day (gpd). Well #3 typically produces approximately 600 to 800 gallons per minute (gpm) when
it is operating, and Well #4 typically produces  approximately 1,200 to 1,600 gpm. According to
the Town of Bernalillo, both  wells pump water from the Rio Grande group aquifer. Well #3 is
approximately 660 feet deep and Well #4 is approximately 970 feet deep.

The supply water for the test is provided from the storage tanks at Well #3, and includes a blend
of water from the two wells. The ARS CFU-50 APC unit was located on the grounds of the
Well #3 site.  The site was secured with a  fence and locked gate, and provided ample space for
adding the piping needed for the test unit and for storage of basic supplies and equipment needed
during the testing.

1.3.2   Source/Feed Water Quality

Table 1-1 presents raw water quality for samples taken from samples collected and analyzed by
ARS between January and April 2006, when the site was evaluated. The water had total hardness

-------
of approximately  70-90 milligrams  per  liter  (mg/L)  as  CaCOs  and the  pH  is  normally
approximately 7.3, based on data collected between June 2002 and March 2004. Water quality
data show that total arsenic concentration varies between 14 and 68 micrograms per liter (|ig/L).
The predominant arsenic species is arsenic (V).

Table 1-1. Raw Water Quality Data
                                                    Concentration
                          Parameter	Units	Range
Total arsenic
Total aluminum
Total iron
Total manganese
Total magnesium
Total calcium
|ig/L
|ig/L
mg/L
Hg/L
mg/L
mg/L
14
<1
0.25
<1
9.7
71
-68
-4
-0.46
-7
-12
-86
1.3.3   Test Site Description

Structural

The ARS CFU-50 APC  system was housed in an 8 foot by 20 foot shipping container. The
containerized system is located next to the water supply building.  The water supply from the
pressurized main system storage tank was piped to the treatment unit. This test site provided the
following advantages:

•   Full electrical supply;
•   Building enclosing the wells and pressure holding tank;
•   Ease of accessibility; and
•   All  required utilities,  including raw water supply, power, and drain  locations for the
    discharge of the filtrate and backwash water to the sanitary sewer system.

Handling of Filtrate

The ARS CFU-50 APC does not have separate discharge ports for backwash or overflow. Water
used for backwash is filtered through a filter press and returned to the reaction  vessel for re-
treatment. For the purposes of this study, all treated water (filtrate) was discharged to one of the
potable water storage tanks maintained by the Town of Bernalillo.

Handling of Residuals

Residual solids are removed from the backwash water with a filter press.  Residual  solids were
stored in 55-gallon drums on-site prior to disposal.

-------
Discharge Permits

No special discharge permits were required for the discharge of the filtrate, and backwash water
from the test unit is recycled back to the reaction tank. The filter cake was characterized as part
of the study (see Section 4.6).  Previous tests conducted by the  vendor indicate this material is
non-hazardous.

-------
                                    Chapter 2
                        Equipment Capabilities and Description

The equipment capabilities and description provided in this section were provided by the vendor
and does not represent verified information.  The ETV evaluation focused on the ability of the
device to remove arsenic from drinking water. Claims beyond arsenic removal are made by the
vendor but were not verified as part of this study.

2.1   Description of Equipment

The ARS CFU-50 APC is a standard,  full-scale,  modular system supplied by ARS  for the
removal  of  arsenic and other  contaminants from  water. The ARS  CFU-50 APC is  a self-
contained, complete  system that connects to a water  supply source.  If the  source is not
pressurized,  a pump, supplied with the unit, is used to pump the water through the treatment
system.  The ARS CFU-50 APC requires a three-phase 480-volt AC  electric power source to
operate the reaction vessel, programmable logic controller (PLC), and  ancillary equipment. The
system used for this test is designed to treat flows up to a maximum flow rate of approximately
35 gpm  (50,000 gpd), from either  a pressurized  or unpressurized water source. Additional
information  on the equipment  installation requirements and  operation of the equipment is
provided in the O&M manual, included Appendix A.

The ARS CFU-50 APC treatment system is fully automated and programmed to control all
aspects of the treatment and filter operation. The control system automatically initiates backwash
cycles based on an inlet pressure level set by the operator. The backwash cycle time is dependent
on the water quality conditions and the amount of solids generated in  the electroflocculation
process.  The control  system monitors data from  the system  operation. This information is
available to  the on-site operator.  Although the system is designed for automatic, unattended
operation, the following information is available to an on-site operator:

          •  Pressure at key points of the device;
          •  Flow rates and throughput totals;
          •  Sand filter data:   regeneration interval, total in-process  times,  current status (on-
             line, back  flushing, standby, etc.);
          •  Electrical process parameters (current and voltage);
          •  Fault/alarm conditions,  based on vessel  levels,  flow rate, pressure levels, gas
             levels, air pressure loss, etc.;
          •  Maintenance messages (for example, filter press cleanout required); and
          •  Oxygen and hydrogen monitor readings.

2.2   Engineering and Scientific Concepts

The ARS CFU-50  APC treatment  system  relies  on electroflocculation which develops an
aluminum flocculent similar to alum and ferric flocculants.  The ARS CFU-50 APC flocculent
generates various hydroxyl water complexes that combine with cations and other contaminants
within the source water.  ARS claims there are several significant differences between the ARS
process and the chemical processes:

-------
   •   The ARS flocculent is generated without the addition of any chemical agents (the anode
       plate is the source of the aluminum used in the flocculation process);
   •   The ARS flocculent does not require any pretreatment or post treatment;
   •   The ARS flocculent does not begin with a salt molecule and therefore does not affect a
       change on water salinity;
   •   The ARS process works in a pH range of 4.5 to 8.5. Higher pH ranges can be reduced
       through a non-chemical ARS method; and
   •   The ARS flocculent particles are a fraction of the size generated through chemical means,
       resulting in floes with extremely high surface area to volume ratios, making the ARS
       process more effective in removing arsenic.

These claims were not verified  as part of the ETV  study.

2.2.1   Physicochemical Efficient Mechanisms

The following two processes are running simultaneously during electrolytic water treatment:

   •   Electrolytic decomposition of water, and
   •   Dissolution of the anodes accompanied by  the formation of metal polyhydroxides and
       metal water complexes.

The main advantage  of  electrolytically-formed flocculent is their adsorbing power.  In this
respect, they are  highly active and show a very good binding capacity for divalent metallic ions.
According to the manufacturer, the ARS process has delivered excellent results for the treatment
of galvanizing wastewaters, dying backwaters, grinding wastewaters, lye solutions, emulsions,
tannery backwaters and similar wastewaters.

Electrolytic  water  decomposition contributes  considerably to the  efficiency  of  complex
procedures.  Hydrogen and oxygen are released in a sequence  of complex mechanisms.  This so-
called nascent hydrogen or oxygen offers a very high potential of reduction and oxidation, which
provides for numerous secondary reactions with the water contents.

2.3    Description of Treatment Train and Unit Processes

With ARS,  a  floe of mixed oxide containing the arsenic contaminant is formed without the
addition of chemicals. Flocculation is  accomplished in a single reaction process, removing heavy
metals.  Water with minimum  electrolytic conductivity is treatable in the reaction tank. Water
with high saline content is managed by regulating the process current level.

Figure 2-1 is a schematic  of the primary components in the ARS CFU-50 APC treatment system.
In the  switch  cabinet (E), all  processes are controlled and monitored.  The power supply (P)
converts the AC  electric current to  a regulated fixed DC current. Untreated/contaminated  water
enters  the  unit through a  regulated  influent  pipe (1).  The  flocculent  generation and
decontamination  process  occurs in the reaction  tank  in a continuous process (2). Flocculent
particles in  the  holding  pipe/tank (3) are subject  to further growth  and reaction  after the

-------
electrolytic process. The water and floe combination is pumped from the reaction vessel, through
the floe pipes, to the sand filters with the filter influent pump.  This pump operates continuously
while the device is in operation.  Filters (4) separate the flocculent from the treated water. The
filter surfaces are cleaned by automatic backwashing, and the flocculation sludge is flushed into
the floe water reservoir tank (5). The low volume, thickened flocculation sludge accumulated in
the floe water reservoir tank is pumped into the filter press (7) by a pump where it is pressed into
a filter cake by a filter  press.  The filter cake must be manually removed from the filter press.
After the treated water passes through the filter, it is stored in the clean water tank (6) for later
use in filter backwashing and rinsing. As the clean water tank level reaches its maximum level, it
is pumped out of the unit through the filtrate water pipe (8).
Figure 2-1. ARS CFU-50 APC schematic view.
The backwash cycle is triggered by an increase in influent pressure across the operating filter
module.  The pressure trigger  for backwash cycles  is set based on local  requirements and
operating characteristics at the site.  The cycle is set based on experience at a site and is typically
set to ensure that at a filter module is backwashed at least once every two days.  The backwash
and rinse  cycle uses treated water for the backwash water source.  Backwash and rinse is
accomplished by pumping treated water at a rate of approximately 100 gpm (14 gpm per square
foot of filter surface area) through the filter module. Backwash is accomplished in an up flow
mode, expanding the granular media bed, and flushing the solids from the media.  Rinse is
accomplished in a downward flow mode, compressing the granular media bed, and flushing the
solids from the media. Approximately 250 to 300 gallons of water is used  for each five-minute
backwash/rinse cycle.  Backwash water from the test system is collected in a waste tank to allow
later dewatering.  During the dewatering process, this water is discharged  back to the reaction
vessel, resulting in zero water loss.

For the ETV test, the feed water was obtained from the water storage tanks  at the Bernalillo test
site.  The ARS system was equipped with a pump to draw water from the tank into the ARS

-------
system.  A pressure regulator and a flow control valve was installed downstream of a double
back flow preventer to control the flow rate of feed water to the system. A flow meter was used
to monitor the flow rate and total flow of feed water to the treatment portion of the process.

A summary of standard operating conditions is provided in Table 2-1  and the ARS CFU-50 APC
system specifications are provided in Table 2-2. Additional equipment  information is provided
in the O&M manual (Appendix A). Figures 2-2 and 2-3 show a schematic and photograph of a
typical system.
Table 2-1.  Test System Operating Conditions
                      Parameter
                   Specification
    Filtrate flow rate
    Backwash flow rate
    Backwash flow velocity
    Backwash water per cycle
    Pressure maximum for backwash initiation
    Feed water pressure
    Source water pressure	
           35 gpm (50,000 gpd)
           lOOgpm
           14 gpm/square foot (ft2)
           250 - 300 gallons
           15 pounds per square inch (psi)
           >20 psi
           >14psi	
Table 2-2. ARS CFU-50 APC System Specifications
    Manufacturer
    Model
    Reactor tank dimensions
    Filter area per module
    Filter module diameter
    Media depth
    Number of filter modules
    Filter pressure rating
    Media per filter module
       Effective size
       Uniformity coefficient
    Skid
    Piping	
ARS USA, LLC
ARS CFU-50 APC
48 in. outer diameter, 48 in. tall
7.1ft2
36 inches
29 inches
2 (alternating in operation)
100 psi max operating pressure
Single media #20 silica sand
0.47 millimeter (mm)
1.42
8 ft x 20 ft shipping container
Schedule 80 PVC
                                          10

-------
                                  -Reactor Plates
   Water IN pipe
 Reactor Tank
   Back wash Pump



          Clean water tank




              Pressure tank
                                                          Clean Floe pipe
                                                                 Clean water pump
                                                                             Top Manifiold
                                                                        Backwash rinse pipe
                                                                      Bottom Manifold
                  Back wash tank
Figure 2-2.  ARS CFU-50 APC right isometric view.
Figure 2-3.  ARS CFU-50 APC skid-mounted unit photograph.
                                           11

-------
2.4    Description of Physical Construction and Components

The ARS CFU-50 APC system is a skid mounted, self-contained unit. The granular media filter
modules are steel tanks with inlet flow distributors, media support plates, and associated fittings,
valves, and piping. The maximum operating pressure is approximately 40 psi. The standard unit
is 20 ft (L) x 8 ft (W) x 8.75 ft (H).  The main system components are (refer to Figure 2-1 for a
schematic view of the components):

1.      Influent water plumbing - To control and regulate influent water flow.
2.      Reaction vessel, which consists of the following components:
          •  Reaction tank - Polyethylene tank for containing electroflocculation equipment.
          •  Reaction frame - Polyethylene frame for holding reaction plates.
          •  Spacer plates - Polyethylene spacers to maintain plate alignment.
          •  Anode plates - Aluminum.
          •  Cathode plates - Graphite.
          •  Level  sensor - To monitor and control the tank water level.
3.      Floe water plumbing - Conveys treated slurry of floe and water from the reaction tank to
       the filters.  The filter influent pump pumps the water through the plumbing, and operates
       continuously when the device is in operation.  The floe water plumbing consists of six-
       inch diameter serpentine pipe, which provides approximately 90 seconds of water and
       floe contact time.
4.      Filter and  filter manifolds - 36-inch diameter, 29-inch deep  single media sand filters,
       used one at a time. One filter is staged and ready for use as back pressure builds in the
       other filter. The filter also has a control manifold on the top and  the bottom of the filters
       to facilitate backwashing and rinsing.
5.      Flocculent tank - A 500-gallon holding tank to temporarily store  the  wastewater
       generated from the backwash/rinse cycles. Water from this tank is transferred to the filter
       press (item 7) to remove the accumulated solids.
6.      Clean water tank and plumbing - A 500-gallon tank to  store treated water for use in back
       washing and rinsing.
7.      Filter press and plumbing - A plumbing system to force water through a plate and frame
       filter press to  dewater the floe from the treatment process. After separation, the remaining
       water is then pumped back to the reaction vessel for recirculation.
8.      Effluent plumbing - To discharge the treated water from the system.
E.      Electrical switch cabinet.
P.      Power supply.

Additional specifications and information are provided in the O&M manual (Appendix A).

2.5    Chemical Consumption and Production of Waste Material

2.5.1   Chemical Consumption

The ARS CFU-50 APC uses the aluminum from the anodes to create a flocculent.  There are no
additional chemicals added or consumed in the process.
                                           12

-------
2.5.2   Waste Production and Physical and Chemical Nature of Wastes

The waste material from the ARS CFU-50 APC is limited to a small amount of hydrogen gas and
filter cake, consisting of the flocculated materials.

Pages  61-65  of the O&M manual provide information on the filter press and its required
maintenance (cleaning).  The filter cake was stored on-site, pending characterization and disposal
at the end of the testing.  Water removed during filtration and filter cake production was pumped
back into the reaction vessel for recirculation.

Some hydrogen is released to the atmosphere by the electrolysis process. Ventilation devices are
built into the system as is a hydrogen monitor to insure that hydrogen concentration remains well
below the lower explosive limit (four percent).  Additional ventilation is provided if hydrogen
buildup of one percent is detected, and if this does not mitigate the situation, an alarm  state is
entered when the level reaches two percent (one-half of the lower explosive limit), automatically
stopping the process. ARS claims that no hydrogen buildup (to even one percent) has ever been
observed except when ventilation was disabled to test the monitor and control logic.

Hydrogen is not classified  as  an atmospheric pollutant. The ventilation equipment  dilutes the
hydrogen with a sufficient quantity of air so that measurement of the resulting output is within
the error band of the monitoring instrument.  Insufficient hydrogen is generated to make capture
for use as a possible fuel a viable option.

ARS also claims that an immeasurable quantity of oxygen is released as free gas. Most of the
oxygen resulting from electrolysis is utilized in  oxidizing reactions associated with the  floe
formation.

2.6    Licensing Requirements

There are no special licensing requirements to operate the ARS CFU-50 APC equipment during
the ETV test.

2.7    Statement of Performance Objectives

The statement of performance objective tested in the verification is:

The ARS CFU-50  APC process is capable of reducing arsenic concentrations from a water
source flowing at a maximum of 35 gpm with a total  arsenic concentration of approximately 14
to 68  |ig/L and a pH of approximately 7.3 to maintain an effluent arsenic concentration less than
10 |ig/L after treatment.

Sampling and analysis of the test site indicated that arsenic concentrations in the 14 to 68 |ig/L
range would be achieved during the verification test.  However, during the verification test, the
arsenic concentrations in the feed water  ranged only from  11 to 14  |ig/L.  An evaluation of the
analytical  data and the test  site  could not identify a  cause  for  this  decrease  in  arsenic
concentrations.  This is discussed in greater detail in Chapter 4.
                                           13

-------
2.8    Advantages of the ARS CFU-50 APC Process

According to ARS, the main advantages of the ARS CFU-50 APC process for removing arsenic
from water are:

•  The process does not require the addition of any water treatment chemicals;
•  The process is flexible and adaptable to the degree of impurities in the source water;
•  The process operates over a wide pH range;
•  The flocculent created during the electrolytic flocculation are easily settleable;
•  The electrolytic flocculation process creates nascent hydrogen and oxygen, which can also
   treat organic compounds, and remove unwanted odors;
•  The electrolytic flocculation process can also remove variety of metals and radiological
   elements (Hg, Pb, Cr, Zn, Cd, Mo, Ni, Ur,  etc.); this claim is outside the protocol and was not
   verified during this test; and
•  The electrolytic flocculation process can also remove a variety of polar and cleavable
   chemicals (not verified during this testing).

The verification testing  did  not include an  evaluation of all  of the  aforementioned vendor
performance claims.

2.9    Potential Limitations of the Equipment

Potential limitations  of the ARS CFU-50 APC process  for the treatment of raw drinking water
with respect to source water quality are (note: these limitations were not verified as part of the
verification test):

•  Poor  water quality in source water can cause high solids loadings to the filter, increasing
   backwash frequency and quantity of solids generated;
•  While the  system is automated and operation should be easy, a  moderate level of operator
   skill may be required for successful use of the system. Variable source water quality  may
   require adjustment of the power setting in order to maintain optimal removal efficiency;
•  Anodes need replenishment to ensure adequate flocculent generation;
•  Possible passivation of the plates over days or weeks as a result of insulation buildup on the
   anode and cathode plates,  depending on the mineral  content of the water.  This may increase
   maintenance requirements;
•  Electrical power consumption settings need to be calibrated to account for source water with
   high salinity; and
•  For source water with fluctuating target  contaminant concentrations, the electrical power
   consumption settings need to be set to  target the highest contaminant concentration; when
   target contaminant concentrations are at the lower end of the range, treatment will still occur,
   however, the higher power consumption setting will increase operating costs.
                                           14

-------
                                     Chapter 3
                               Methods and Procedures

The testing methods and procedures were specified in the Product Specific  Test Plan for the
Advanced Remediation Systems USA, LLC ARS CFU-50 APC Electroflocculation and Filtration
Water Treatment System for Arsenic Removal from Drinking Water (NSF International, March
2006).   The PSTP, included in Appendix B,  is summarized in this section.  Deviations to the
PSTP are summarized in Section 4.8 of this report.

3.1    Quantitative and Qualitative Evaluation Criteria

As defined in the ETV protocol, the objectives of the verification are to evaluate equipment in
the following areas:

•   Report the  actual results obtained by the equipment as operated under the conditions at the
    test site;
•   The measurement of residual materials generated during testing;
•   The impacts on performance of any variations in feed water quality or process variation;
•   The logistical, human and other resources necessary to operate the equipment; and
•   The reliability, ruggedness, ranges of usefulness and ease of operation of the equipment.

3.2    Key Water Quality Parameters

Key water quality parameters used for  evaluation of the ARS  CFU-50  APC are listed in
Table 3-1.  The Water Quality  and  Inorganic Parameter columns are the key  parameters for
evaluating  the treatment process and water quality.   The  parameters listed in the Other
Parameters column should not have an  immediate impact on the treatment process, but are
important parameters in drinking water supplies.

Table 3-1.  Key Filtrate Water Quality Parameters
Water Quality
• Temperature
• Alkalinity
• Hardness
• pH
• Turbidity
• Residual Chlorine


Inorganic Parameters
• Arsenic (speciation) «
• Iron «
• Aluminum «
• Total suspended solids
(TSS) «
«
«
«
Other Parameters
• Manganese
• True Color
• Total Organic Carbon
(TOC)
• Chloride
• Sulfate
• Fluoride
• Dissolved Oxygen (DO)
3.3    Operations and Maintenance

ARS provided a draft O&M manual with the ARS CFU-50 APC, which is included in Appendix
A.  As part of the verification testing, the ETV DWS Center reviewed the O&M documentation
                                          15

-------
for the ARS CFU-50 APC.  Results of the review are included in this ETV report. In addition,
the following aspects of operability are addressed in the report:

•  Fluctuation of flow rates and pressures through the unit;
•  Presence of devices to aid the operator with flow control adjustment;
•  Availability of pressure measurement;
•  Measurement of feed water rate of flow;
•  Adequacy and ease of use of the PLC control system;
•  Ease of operating the computer control system;
•  Generation of residual materials; and
•  Availability of process data to the operator.

3.4    Environmental Technology Verification Testing Plan

The PSTP for the verification test was prepared in accordance with the ETV Protocol.  The PSTP
divided the work into three main tasks (A, B, C) with Task C, the verification test itself, divided
into six tasks. The PSTP included a Quality Assurance Project Plan (QAPP), which specified
procedures to be used to ensure the accurate  documentation of both water quality and  equipment
performance.

An overview of each task is  provided below with detailed  information on testing procedures
presented in later sections.

3.4.1   Task A: Raw Water Characterization

The objective of Task A was to obtain a chemical and physical characterization of the  raw water.
Information on the groundwater supply that provides the  raw water was needed to aid in
interpretation of feed water characterization.  Grab samples of the raw water were analyzed for
the parameters indicated in Table 3-1.

3.4.2   Task B: Arsenic Loss Test

During Task B,  The ARS  CFU-50 APC was run without  supplying  electrical power to the
reaction vessel to  evaluate the arsenic loss across the treatment train without powering the
electroflocculation process.

The system was flushed to remove treated  water from the  tanks and  piping, the filters were
backwashed, and the  waste material was removed.  Following system clean out, the system was
operated continuously for 24 hours.  Feed water and filtrate samples were collected at six-hour
intervals and analyzed for the parameters indicated in Table 3-1, in accordance with the PSTP.

3.4.3   Task C: Verification Test Procedures

Task 1: Verification Testing Runs

The ARS  CFU-50  APC was operated over a 14-day timeframe to collect  data on  equipment
performance and water quality for purposes of performance verification. During  this timeframe,
                                           16

-------
operational problems with the filter press caused the system to shut down, resulting in an actual
operation time of 287 hours, less than identified in the PSTP.   The operational problems are
described  in  greater detail in  Chapter 4.  Daily measurements and observation of operating
parameters were made, and samples collected of the feed water and filtrate for analysis. Testing
included one 48-hour intensive survey period. Results are presented in Chapter 4.

Task 2: Raw Water, Feed Water, and Filtrate Water Quality

During verification testing, feed water and filtrate water samples were collected and appropriate
sample  analyses  performed.     Samples  were   analyzed  for  aluminum to  monitor the
electroflocculation process, arsenic to evaluate arsenic removal, and other water quality analyses,
such as pH, turbidity, hardness, alkalinity, etc., to monitor the impact of the treatment process on
water quality.

Task 3: Operating Conditions and Performance

During  verification testing, operating  conditions and  performance  of the water  treatment
equipment  were documented.  Equipment performance information collected included data on
filtrate flow rate and total filtrate volume produced,  pressure differential across the granular
media filters,  electrical energy used and maintenance required during operation.

The  PSTP called  for collection  of  other filter operation data,  including filter run lengths,
frequency and duration  of backwash cycles, and volume of water treated per filter run.  This
information was not collected during the testing, as described in Chapter 4.

Task 4: Total Arsenic Removal

Total arsenic in  the feed and filtrate samples were measured to  evaluate  total arsenic removal
during verification testing.  Samples were collected daily  over the 14-day period.  This test phase
included a 48-hour intensive sampling period that occurred at the end of the first week of testing.
During this phase, samples were collected at the start (hour 0) and after hours 1, 3, 6, 12, 18 and
24; the filter was then backwashed and samples were  collected at the same time intervals over
the next 24 hours as during the first 24 hours.

All samples were analyzed for total arsenic, aluminum,  pH, iron and residual chlorine.  Other
water quality parameters were analyzed  at less frequent intervals.  Speciation of arsenic was
completed on samples collected at hours 0, 24 and 48 of the intensive sampling period.

Task 5: Data Management

The objective of this task was to establish an effective  field protocol for data management at the
field operations  site, and  for data transmission between the FTO  and the ETV  DWS Center.
Master field logs were  prepared  and field sheets  for data collection were used to ensure  all
scheduled activities were performed.  The logs were delivered to the ETV DWS Center project
coordinator on a weekly basis.
                                           17

-------
Task 6: Quality Assurance/Quality Control (QA/QC)

An important aspect of verification testing was the development of specific QA/QC procedures.
The objective of this task was to assure accurate measurement of operational and water quality
parameters during the verification test.   Weekly and  one-time QA/QC verifications  were
specified in the QAPP in Chapter 5 of the PSTP.  Equipment flow rates were documented on a
daily basis, and a daily walkthrough was completed to verify that each piece of equipment or
instrumentation was operating properly.  An audit of the FTO was also  conducted during the
testing.

3.5    Operation and Maintenance

An O&M manual was  received from ARS  when the ARS CFU-50 APC was installed.  NSF
reviewed the O&M manual and evaluated the instructions and procedures for their applicability
during the verification test and for overall completeness.

3.5.1   Operability Evaluation

The basis of the review and evaluation for equipment operability during verification testing was
formed from the factors listed below. These aspects of plant operation are  reported, to the extent
practical, in Chapter 4 of this report.

The factors considered included:

   •   Can automatic backwash be initiated by:
       -  Reaching a set value for head loss?
       -  Default minimum time?
   •   Is granular media pressure differential measurement provided?
   •   Is rate of flow of feed water measured?
   •   Is backwash rate of flow measured and variable?
   •   Is backwash duration (time) variable?

Other factors and questions included:

   •   Does the equipment have sensors or monitoring equipment that can detect an equipment
       malfunction, unsatisfactory filtrate water  quality, or  operating conditions that exceed
       allowable limits?
   •   If so, during such situations can the equipment be automatically shut down?
   •   Upon automatic shutdown, can notification be provided if the operator is not present?
                                           18

-------
                                      Chapter 4
                                 Results and Discussion
4.1    Introduction
The verification test program for the ARS CFU-50 APC began with equipment installation at the
Bernalillo Well #3 site in Bernalillo, New Mexico, in April 2006 and ended with the completion
of the verification test on May  2, 2006. The test site was described in Section 1.3.  The ARS
CFU-50 APC was described in Chapter 2.

The equipment was installed prior to the beginning of the ETV tests.  Raw water characterization
samples were collected on February 24 and March 9, 2005, prior to ETV tests.  The arsenic loss
test was  performed from May  1-2, 2006.  The 14-day  verification test, including a  48-hour
intensive survey, was performed from April 18 through May 1, 2006.

This chapter presents a  summary of the water quality and operating data collected during the
verification test.   Activities  and  data  collected  during  the start-up  and shakedown of the
equipment, and  the  raw water  characterization were performed prior  to  the  actual  14-day
verification test.   The arsenic loss test was performed immediately after the 14-day verification
test.  The results from the 14-day verification test are presented, including data on the feed and
filtrate water arsenic concentration and  other water quality parameters.  Operating data are
presented to describe the flow rates, volume of treated water produced, backwash information,
pressure  differential across  the sand filter, electrical power, and related operating information.
QA/QC information,  as described by the QAPP in the PSTP for this verification test, is presented
at the end of the chapter.

4.2    Equipment Installation, Start-up, and Shakedown

At the beginning of the ETV project, ARS and FTO personnel performed a thorough evaluation
of the installation.  This included ARS training FTO personnel on operations, maintenance of the
device for FTO personnel, and FTO personnel conducting an evaluation on such things as how
and where water samples would be collected, where critical flow and pressure readings would be
read and recorded,   a  full  evaluation  of the PLC's  operating  capabilities,   maintenance
requirements (especially how to maintain the filter press),  emergency/safety considerations, and
startup/shutdown  operations). Based on tests conducted by ARS prior to ETV testing, it was
determined that 30 amps of electrical power would need to be delivered to the reaction vessel in
order to reduce the arsenic concentrations to a level consistently below 10 |ig/L.

4.2.1   Flow Measurement

As part of normal operating conditions, the feed and filtrate water pumps, which pump water into
the reaction vessel and  drinking water reservoir tank, respectively, shut off intermittently,  as
controlled by the PLC. A high water level sensor in the reaction vessel would shut off the feed
water flow when actuated, and the high and low level sensors in the drinking water reservoir tank
would  actuate the filtrate water  pump.  The instantaneous flow rate readings noted by the FTO
were recorded when the respective pumps were operating.  The actual  flow rate through the
system is less than either of the readings from  these flow monitors.  For the purposes of this
                                           19

-------
verification, the average flow through the system was calculated by dividing the total volume of
treated water for each test, which was recorded by the totalizer;  and by the total operating time,
which was recorded by the PLC or by FTO personnel.

4.3    Task A:  Raw Water Characterization

ARS and the Town of Bernalillo characterized the raw water prior to the start of ETV testing.
This characterization demonstrated that the raw water posed a challenge sufficient to verify the
performance of the ARS CFU-50 APC, while not creating conditions that were disadvantageous
to the  device's performance.  Since the  data were  not collected by  an ETV-approved testing
organization, it was not included as part of the ETV study.

Samples of the feed water from the combined water tanks were  collected on the first day of the
verification testing and are used for the "raw" water characterization. The data for these samples
are presented in Table 4-1.

Table 4-1. Feed Water Characterization Data - April 18, 2006
Parameter
pH
Temperature
Turbidity (bench top)
Alkalinity
Free Chlorine
Total Chlorine
DO
True Color
Total Arsenic
Dissolved Arsenic
Arsenic (III)
Arsenic (V)
Iron
Aluminum
Manganese
Chloride
Sulfate
TOC
Fluoride
Calcium
Magnesium
Hardness(3)
Units
Standard Units (S.U.)
°C
Nephelometric Turbidity
Units (NTU)
mg/L CaCO3
mg/L
mg/L
mg/L
Color Units (C.U.)
Mg/L
Mg/L
Mg/L
Mg/L
mg/L
Mg/L
Mg/L
mg/L
mg/L
mg/L
mg/L
mg/L
mg/L
mg/L as CaCO3
Result
7-8(1)
20.2
0.47
130
0.71
0.84
7.40
Q(2)
14
12
20
<2
<0.02
71
<1
180
110
0.3
0.3
74
11
230
              pH samples analyzed with Litmus paper due to instrument malfunction.
          (2)  Analyzed on Day 2.
          (3)  Calculated from calcium and magnesium concentrations.
                                           20

-------
4.4    Task B: Initial Test Runs

4.4.1   Arsenic Loss Test

The arsenic loss test, to determine if arsenic is removed and retained by the system without
electricity supplied to the reaction vessel, was performed over a 24-hour period on May  1-2,
2006, after Task C had been completed.  The PSTP specifies that a shakedown and arsenic  loss
test be run on the verified device prior to the start of the Task C test. For this installation, ARS
had conducted numerous verification runs prior to the start  of verification, and had the system
ready for verification when FTO personnel arrived.  Prior to  the start of the arsenic loss test, the
cathodes and anodes were removed from the reaction vessel, so it was decided to conduct the
Task C test first. The cathodes and anodes were removed after the Task C test was complete.

Prior to the  start of the arsenic loss test, the storage tanks within the system were emptied, the
sand filters were backwashed, and feed water was run through the device for approximately  270
minutes to flush out the flocculent materials.  The automated backwash cycle was also disabled,
so that a single sand filter was challenged for the entire test duration.

The operating data and results from the 24-hour arsenic loss  test are shown in Table 4-2.  Based
on the flow monitor readings, the  feed  flow rate averaged 36.1  gpm and the  filtrate flow  rate
averaged 50.2 gpm.  However, as noted in Section  4.2.1, these flow readings were taken when
the respective  pumps  were  operating,  and they do not always operate  as  part of normal
operations.  The total  volume  processed over the  24-hour  period was 44,755  gallons,  which
results in  a  calculated  average flow rate of 31 gpm.   The filtrate pressure  increased  over the
24-hour period from 7.5  psi to 15.5 psi  as the sand bed became compacted and charged with
contaminants.

Table  4-2. Task B Arsenic Loss Test Operating Data

Day Hour
1 0
6
2 12
18
24
Feed
Pressure
(psi)
18
17
18
16
16
Filtrate
Pressure
(psi)
7.5
11.2
11.5
15.4
15.5
Pressure
Delta(1)
(psi)
10.5
5.8
6.5
0.6
0.5
Total Volume
Treated
(gal)
0
10,779
24,590
35,236
44,755
Flow Rate(2)
(gpm)
~
29.9
34.2
32.6
31.1
   (1) Pressure Delta is the pressure differential or head loss through the filter as measured by the pressure
       difference between the feed and filtrate.
   (2) Flow rate is calculated by dividing the total volume treated by 60 times the hour.
                                           21

-------
Table 4-3 presents the water quality for the arsenic loss test.  The statistical calculations of these
data are  presented in Appendix C.  There was no loss of arsenic through the system over the
24-hour test, with both the feed and filtrate water total arsenic averaging 11 |ig/L.  Arsenic (III)
was the  predominant arsenic species  in the feed water.  Aluminum concentrations  increased
slightly through  the system.   All  other water quality indicators remained  steady and passed
through the filter.
                                            22

-------
Table 4-3. Task B Arsenic Loss Test Water Quality Results
Parameter
pH
Temperature
Turbidity (bench top)
Alkalinity
Free Residual Chlorine
Total Residual Chlorine
True Color
Calcium
Magnesium
Hardness0 -1
Total Arsenic
Dissolved Arsenic
Arsenic (III)
Arsenic (V)
Iron
Aluminum
Manganese
Chloride
Sulfate
Fluoride
TOC
DO
Units
S.U.
°c
NTU
mg/L as CaCO3
mg/L
mg/L
C.U.
mg/L
mg/L
mg/L
Mg/L
Mg/L
Mg/L
Mg/L
mg/L
Mg/L
Mg/L
mg/L
mg/L
mg/L
mg/L
mg/L
Feed Water
0 hours 6 hours 12 hours 18 hours
7.6 7.7 7.7 7.7
23.7 22.7 21.7 23.7
0.28 0.20 0.17 0.31
140 140 140
0.58 0.57 0.51
0.59 0.60 0.58
-
-
-
-
10 13 11
11
9
2
<0.02 <0.02 <0.02
<10 <10 23
.
170
110
.
-
7.09 6.95 6.78 7.31
24 hours
7.8
24.1
0.21
93
0.52
0.55
1.0
93
11
280
9
11
6
5
<0.02
27
<1
170
110
0.3
<0.1
7.27
0 hours 6 hours
7.8 7.8
24.0 23.0
0.17 0.30
160
0.58
0.60
-
-
-

11
10
11
<2
<0.02
33
-
170
110
-
-
6.80 6.85
Filtrate
12 hours 18 hours 24 hours
7.8 7.8 7.8
22.2 23.7 24.5
0.21 0.19 0.14
140 140 140
0.58 0.55 0.53
0.64 0.61 0.58
2.0
90
12
270
11 12 11
11
6
5
<0.02 <0.02 <0.02
27 25 28
<1
170
100
0.3
<0.1
6.66 7.02 6.74
    Calculated from calcium and magnesium concentrations.
                                                               23

-------
4.5    Task C: Verification Test

4.5.1   Operating Results

The ARS CFU-50 APC was set to the operating criteria established by ARS prior to ETV testing.
Electrical  power settings to the reaction vessel and other operational settings were  set and
verified prior to the start of testing.  The verification test was started on April 18, 2006.

Table 4-4  shows the daily operating data for the verification test.  During the test, there were a
total of four incidents (April 20, 21, 28, and 30) where a sensor triggered the PLC to shut down
operations. During each incident, the sensor indicated that either the floe water reservoir tank
(see Figure 2-1) had exceeded capacity or  the filter  press alarm went off.  Under the FTO's
supervision, ARS personnel analyzed the incidents and  determined that in each instance, the
filter press had clogged to a point where it  was prohibiting sufficient filtration to maintain the
device's rated throughput.   ARS personnel recommended that the filter  press be cleaned a
minimum  of once every 24 hours to prevent the ARS CFU-50 APC from automatically shutting
down.   After  each shutdown incident, FTO  personnel  cleaned the filter press and  resumed
operation  in accordance with the startup procedures outlined in the ARS O&M manual.  As a
result of  these incidents, the ARS CFU-50  APC experienced approximately  36 hours of
downtime during the 14-day verification test.

The flow rate noted in Table 4-4 was calculated by dividing the volume of water treated by the
device by the operating time.   It is not associated with the instantaneous readings of any
particular  pump. During days when the device was functioning properly, the typical volume of
water produced ranged  from approximately 45,000 to 47,000 gpd or approximately 31.3 to 32.6
gpm.  During  these days, the average flow rate was  32.1  gpm.  Over the entire test duration
(approximately 287 hours), the filtrate flow rate was approximately 30.7 gpm. The filter influent
pump was programmed to operate at a near-constant flow rate of 35 gpm. The difference in the
flow rate for the device as  a whole and this pump can be attributed to recirculated backwash
water.

The  system pressure was monitored  at the feed and filtrate  water locations  (upstream and
downstream of the sand filters).  The FTO technician recorded the pressure readings in Table 4-4
manually as part of routine sampling and inspections.  The ARS CFU-50 APC backwash cycles
were programmed to initiate when the pressure differential reached 15  psi.  The FTO pressure
readings were not scheduled to  evaluate  whether the  pressure differential reached  15  psi.
Furthermore, the ARS CFU-50 APC PLC was not programmed to record pressure differentials at
the start of backwash cycles, so the pressure  differential  evaluation for this verification was
limited to  whether the differential exceeded  15  psi  during the time the FTO personnel inspected
the device. There was one instance (April 28)  when the  pressure differential reached  15.9 psi;
otherwise, the pressure  differential noted by  FTO personnel during sampling and inspection was
below the  15 psi threshold, and averaged 10.9 psi during the inspections.

The amperage to the reaction vessel remained  constant at 30 amps throughout the verification
test, while the voltage averaged 4.23 volts and ranged between 3.84 and 5.51 volts.
                                           24

-------
Table 4-4. Operating Data


Date
4/18/06
4/19/06
4/20/06
4/21/06
4/22/06
4/23/06
4/24/06
4/25/06
4/26/06
4/27/06
4/28/06
4/29/06
4/30/06
5/01/06
Number of
samples
Average
Maximum
Minimum
Std.
Deviation
95% Conf.
Interval
Total
Filtrate
Volume
(gal)
0
44,394
91,380
110,415
158,678
204,385
249,651
297,422
343,835
389,460
428,847
463,324
482,300
527,779
NC
NC
NC
NC
NC
NC

Flow
Rate(1)
(gpm)
~
30.8
29.3
29.7
30.8
31.0
31.1
31.4
31.5
31.5
31.3
30.6
29.1
30.6
13
30.7
31.5
29.1
0.81
30.1-31.2



Pressure (psi)
Feed
18
19
18
18
19
17
18
18
17
18
21
18
18
18
14
18
21
17
0.97
18-19
Filtrate
7.4
10.3
9.3
5.0
9.0
10.6
10.5
7.6
4.0
7.4
5.1
4.1
4.3
7.7
14
7.3
10.6
4
2.4
5.7-9.0
Delta
10.6
8.7
8.7
13.0
10.0
6.4
7.5
10.4
13.0
10.6
15.9
13.9
13.7
10.3
14
10.9
15.9
6.4
2.7
9.1-12.7



Operating
Electrical Power(2) Time
Amps
30
30
30
30
30
30
30
30
30
30
30
30
30
30
14
30
30
30
0
30-30
Volts
NR
NR
3.94
3.84
NR
4.14
3.97
3.98
4.02
4.11
4.06
4.57
4.37
5.51
11
4.23
5.51
3.84
0.47
3.85-4.60
(hours)
16
40
52
62
86
110
134
158
182
206
228
252
276
287
NC
NC
NC
NC
NC
NC
(1)   The flow rate was calculated by dividing the total filtrate volume by the operating time and multiplying the
    quotient by 60 minutes/hour.
(2)   Average of three contactors.
NC = Not calculated.
NR = Reading not recorded.

4.5.2  Arsenic Results

The determination of total arsenic removal  using the ARS CFU-50 APC was the primary
objective of the verification test.  The arsenic results for  the feed and filtrate water monitored
daily during the verification test are presented in this section. Also included are the results from
the 48-hour intensive survey,  when samples  for arsenic analysis were collected on  a more
frequent basis.  The total arsenic data are presented in Tables 4-5 and  4-6.  Arsenic speciation
data are presented in Table 4-7.  Figure 4-1 shows the arsenic results plotted for the  14-day
verification test.
                                             25

-------
Table 4-5. Daily Total Arsenic Results (ug/L)
Date
4/18/06
4/19/06
4/20/06
4/21/06
4/22/06
4/23/06
4/24/06
4/25/06
4/26/06
4/27/06
4/28/06
4/29/06
4/30/06
5/01/06
Number of samples
Average
Maximum
Minimum
Std. Deviation
95% Conf. Interval
Feed
14
11
11
13
11
11
11
12
11
11
12
11
12
12
14
12
14
11
0.9
(11-12)
Filtrate
6
5
6
5
4
5
6
6
5
6
6
5
7
6
14
6
7
4
0.8
(5-6)
Based on the daily sample results, the total arsenic in the feed water averaged 12 |ig/L.  Over the
14-day period, the maximum total arsenic was 14 |ig/L in the feed water and the minimum was
11 |ig/L.  The arsenic speciation data for the feed water showed that most of the arsenic was
present  as  arsenic (III),  with  some  arsenic  (V) also present.  The  average  total arsenic
concentration in  the  filtrate was 6  |ig/L, with a minimum concentration of 4  |ig/L and a
maximum concentration of 7 |ig/L.

The data collected during the 48-hour intensive  survey were  consistent with the data  collected
each day during the verification test.   There was no  indication  of any transient or short time
changes in the arsenic concentration or other monitored parameters.
                                           26

-------
Table 4-6. Total Arsenic Results for 48-Hour Intensive Survey (ug/L)
Date
4/25/06
4/25/06
4/25/06
4/25/06
4/25/06
4/26/06
4/26/06
4/26/06
4/26/06
4/27/06
4/27/06
Number of samples
Average
Maximum
Minimum
Std. Deviation
95% Conf. Interval
Time (hours)
0
1
3
6
12
18
24
30
36
42
48






Feed
12
12
9
11
11
10
11
10
12
13
11
11
11
13
9
1.1
(10,12)
Filtrate
6
5
6
5
5
5
5
6
6
6
6
11
6
6
5
0.5
(5,6)
                                        27

-------
Table 4-7.  Arsenic Speciation Data (jig/L)
Total Arsenic
Date
4/18/06
4/19/06
4/20/06
4/21/06
4/22/06
4/23/06
4/24/06
4/25/06
4/26/06
4/27/06
4/28/06
4/29/06
4/30/06
5/01/06

Day 1
Day 2
Day 3
Day 4
Day 5
Day 6
Day 7
Day 8
Day 9
Day 10
Day 11
Day 12
Day 13
Day 14
Number of samples
Average
Maximum
Minimum



Std. Deviation
95% Conf.
Interval
Feed
14
11
11
13
11
11
11
12
11
11
12
11
12
12
14
12
14
11
0.9
(11-12)
Filtrate
6
5
6
5
4
5
6
6
5
6
6
5
7
6
14
6
7
4
0.8
(5-6)
Dissolved Arsenic
Feed
12
11
10
12
3
10
10
11
11
10
10
10
11
11
14
10
12
3
2.2
(8-12)
Filtrate
18,000(2)
3
4
-
8
3
4
4
3
4
4
4
6
4
12
4
8
3
1
(2-6)
Arsenic (III)(1)
Feed
20
<1
16
7
<1
2
14
7
15
<1
12
6
10
12
14
9
20
<1
6
(4-14)
Filtrate
8
4
5
16
<1
9
7
1
4
7
<1
9
2
4
14
6
16
<1
4
(1-10)
Arsenic (V) (1)
Feed
<1
10
<1
5
2
8
<1
4
<1
9
<1
4
1
<1
14
4
10
<1
3
(
-------
       16
   _  14
   |>  12
   7T  10
   "c
    a)   8
   I   6
        0
W   \
                                           ^.
                                  Date
Figure 4-1.  Verification test daily arsenic results.

4.5.3   Feed and Filtrate Water Quality Results

Water quality data were collected each day for pH, temperature, turbidity, and chlorine (total and
free residual).  Samples for aluminum and alkalinity analyses were also collected daily. DO was
monitored daily in the feed and filtrate water, as it can affect the oxidation of aluminum and
arsenic (III).   Iron was collected on a  daily basis. Other water quality parameters, including
calcium, magnesium, manganese, sulfate, chloride, fluoride, TOC, and color, were monitored on
a weekly basis.   All of the field data log sheets and NSF laboratory reports are included  in
Appendices D and E.

Tables 4-8 and 4-9 present the individual pH measurements for the daily samples and for the
48-hour intensive survey.  Figure 4-2 shows the pH for the feed and filtrate water from the daily
samples.  During the verification test, the feed water pH was steady in the range of 7.6-7.8, with
a median of 7.7. The filtrate pH was very similar to the feed water pH, as expected. The filtrate
pH ranged from 7.7-7.9, with a median value of 7.7.  The pH during the 48-hour intensive survey
was monitored frequently and displayed similar results to the daily pH levels found over the
14-day verification test.
                                           29

-------
Table 4-8. pH Results (S.U.)
Date Feed
4/18/06 7-8(1)
4/19/06 7.7
4/20/06 7.7
4/21/06 7.7
4/22/06 7.7
4/23/06 7.6
4/24/06 7.8
4/25/06 7.7
4/26/06 7.7
4/27/06 7.7
4/28/06 7.7
4/29/06 7.8
4/30/06 7.7
5/01/06 7.8
Number of samples 1 3
Median 7.7
Maximum 7.8
Minimum 7.6
Filtrate
70)
7.7
7.7
7.8
7.7
7.7
7.8
7.7
7.7
7.7
7.7
7.7
7.9
7.7
13
7.7
7.9
7.7
(1) pH samples analyzed with Litmus paper due to
instrument malfunction; this data was not used in
the statistical calculations.
Table 4-9. pH Results for the 48-Hour Intensive Survey
„ Time „ ,
Date ,u , Feed
(hours)
4/25/06 0 7.7
4/25/06 1 7.7
4/25/06 3 7.7
4/25/06 6 7.7
4/25/06 12 7.7
4/26/06 18 7.8
4/26/06 24 7.7
4/26/06 30 7.7
4/26/06 36 7.7
4/27/06 42 7.7
4/27/06 48 7.7
Number of samples 1 1
Median 7.7
Maximum 7.8
Minimum 7.7

(S.U.)
Filtrate
7.7
7.7
7.6
7.7
7.7
7.8
7.7
7.6
7.7
7.7
7.7
11
7.7
7.8
7.6
                                         30

-------
       7.9 n
       7.8
       7.7 -
       7.6
                        ^
                                                                   ^
                                     Date
Figure 4-2. Verification test pH results.
Tables 4-10 and 4-11 present the individual turbidity measurements for the daily turbidity levels
and for the 48-hour intensive survey.  Figure 4-3  shows the turbidity for the feed and filtrate
water from the daily samples.  The filtrate turbidity was higher than the feed turbidity throughout
the verification test, averaging 0.80 NTU in the filtrate and 0.30 NTU in the feed water. Results
during the 48-hour intensive survey  were very similar to the daily results over the 14-day
verification test, averaging 0.90 NTU in the filtrate and 0.30 in the feed water. The increase in
turbidity  is  likely  attributable  to  the  fine nature  of the  flocculent  formed  during  the
electroflocculation process.

-------
Table 4-10. Bench Top Turbidity Results

Date
4/18/06
4/19/06
4/20/06
4/21/06
4/22/06
4/23/06
4/24/06
4/25/06
4/26/06
4/27/06
4/28/06
4/29/06
4/30/06
5/01/06
Number of samples
Average
Maximum
Minimum
Std. Deviation
95% Conf. Interval
Turbidity
Feed
0.47
0.19
0.23
0.30
0.31
0.30
0.42
0.26
0.37
0.21
0.20
0.30
0.18
0.26
14
0.30
0.45
0.20
0.09
(0.25-0.35)
(NTU)
Filtrate
0.79
0.93
1.06
0.45
0.90
1.21
1.08
1.00
0.55
0.87
0.63
0.49
0.37
1.06
14
0.80
1.2
0.35
0.27
(0.65-1.0)
                                        32

-------
Table 4-11.  Bench Top Turbidity Results for the 48-Hour Intensive Survey

















Bench Top Turbidity (NTU)
ooooo-».-».-».
dk>ji.cr>bodk>ji>.
oooooooo


Date Time
4/25/06
4/25/06
4/25/06
4/25/06
4/25/06
4/26/06
4/26/06
4/26/06
4/26/06
4/27/06
4/27/06
Number of samples
Average
Maximum
Minimum
Std. Deviation
95% Conf. Interval


(hours)
0
1
3
6
12
18
24
30
36
42
48


Turbidity (NTU)
Feed Filtrate
0
0
0
0
0
0
0
0
.26 1.00
.55 1.06
.28 1.00
.26 0.44
.32 1.00
.31 0.85
.37 0.55
.19 0.90
0.22 0.91
0
0

.24 1.19
.21 0.87
11 11
0.30 0.90



0
0
0
.55 1.2
.20 0.45
.10 0.20
(0.20-0.35) (0.70-1.1)



•

S\ !
\ /
^ V
\j* • **
^j^"^
\
\

' V-
A
/
M
^
/
/ 1- _1
\/ — •— Feed
v / -"-Filtrate

vy
^^ ^*
-
i i

\
Date
I

I i

Figure 4-3. Verification test turbidity results.
                                          33

-------
Table 4-12 presents the alkalinity during the verification test. The feed water averaged 130 mg/L
as CaCOs  and was stable  throughout the  test.  The maximum feed water concentration was
140 mg/L and the minimum was 93 mg/L. The filtrate alkalinity also averaged 130 mg/L, with a
maximum of 150 mg/L and a minimum of 130 mg/L.  Figure 4-4 presents the alkalinity results
for the feed and filtrate water during the verification test. The alkalinity concentration during the
48-hour intensive survey was  slightly higher in the  filtrate  (130 mg/L) than in the feed water
(140 mg/L), as shown in Table 4-13.

Table 4-12. Alkalinity Results
           Date
Alkalinity (mg/L as CaCO3)
Feed                Filtrate
4/18/06
4/19/06
4/20/06
4/21/06
4/22/06
4/23/06
4/24/06
4/25/06
4/26/06
4/27/06
4/28/06
4/29/06
4/30/06
5/01/06
Number of samples
Average
Maximum
Minimum
Std. Deviation
95% Conf. Interval
130
140
140
140
140
140
140
140
140
140
140
93*
130
120
14
130
140
93
13
(121-139)
130
130
130
130
130
130
150
130
130
130
140
140
130
140
14
130
150
130
6.3
(130-130)
          NR = Not Recorded.
            = Result considered anomalous, but was used in statistical evaluations.
                                           34

-------
       160 -i
                                    Date
Figure 4-4.  Verification test alkalinity results.
Table 4-13. Alkalinity Results for the 48-Hour Intensive Survey
                Date
 Time
(hours)
Alkalinity (mg/L as CaCO3)
  Feed            Filtrate
4/25/06
4/26/06
4/27/06
0
24
48
140
140
140
130
130
130
Table 4-14 and Figure 4-5 present the total aluminum concentrations measured in the feed and
filtrate water during the verification test.  Aluminum was detected in four of the 14 feed water
samples, at  concentrations ranging  from 13 to 84 |ig/L, while the remaining ten feed water
samples had aluminum concentrations below the 10 |ig/L detection limit. NSF QA conducted an
evaluation  of the four  samples where  aluminum was detected  to evaluate  whether the
concentrations could be the result of laboratory error.  The evaluation yielded no explanation
attributable to the laboratory testing procedures that would indicate a false positive result.  In the
filtrate, the average aluminum concentration was 560 |ig/L, and ranged from 200 to 890 |ig/L.
This average filtrate aluminum concentration is 20  times greater than the feed water average
concentration and significantly higher than the National Secondary Drinking Water Regulation
range of 50 to 200 |ig/L. The electroflocculation process used by  this technology generates
aluminum hydroxide flocculent used in the removal of arsenic. It appears that there is an excess
of aluminum hydroxide generated in the process,  which can not all be removed by the filters,
resulting in  an aluminum concentration in the filtrate above the EPA  secondary regulation for
aluminum (200 |ig/L).
                                           35

-------
Table 4-14. Aluminum Results
Aluminum (ug/L)
Date
4/18/06
4/19/06
4/20/06
4/21/06
4/22/06
4/23/06
4/24/06
4/25/06
4/26/06
4/27/06
4/28/06
4/29/06
4/30/06
5/1/06
Number of samples
Average
Maximum
Minimum
Std. Deviation
95% Conf. Interval
Feed(1)
71
<10
<10
48
<10
<10
79
<10
<10
<10
<10
84
<10
13
14
28
84
<10
29
(8-47)
Filtrate
540
770
700
200
680
870
780
580
310
580
400
340
250
890
14
560
890
200
230
(400-720)
            (1)  Concentrations reported as <10 set equal to the detection limit
                for calculating statistics.
The aluminum concentrations during the 48-hour intensive survey are presented in Table 4-15.
The results during the 48-hour intensive  survey were similar to the results from the 14-day
verification test. During the intensive survey, aluminum was detected in two of the eleven feed
water samples at concentrations of 77 and 79 |ig/L.  The filtrate water had an average aluminum
concentration of 550 |ig/L, and ranged from 220 |ig/L to 740 |ig/L.
                                            36

-------
0)


₯
3
c

E
_3

<
                                                                      •Feed


                                                                      •Filtrate

                                                                       Secondary MCL
                                Date
Figure 4-5.  Verification test aluminum results.
Table 4-15. Aluminum Results for the 48-Hour Intensive Survey

Date
4/25/06
4/25/06
4/25/06
4/25/06
4/25/06
4/26/06
4/26/06
4/26/06
4/26/06
4/27/06
4/27/06
Number of samples
Average
Maximum
Minimum
Std. Deviation
95% Conf. Interval

Time (hours)
0
1
3
6
12
18
24
30
36
42
48






Aluminum (j
Feed(1)
<10
<10
<10
<10
79
77
<10
<10
<10
<10
<10
11
22
79
<10
28
(< 10-44)
Ag/L)
Filtrate
580
600
640
220
610
610
310
570
610
740
580
11
550
740
220
150
(430-670)
              Concentrations reported as <10 set equal to the detection limit for calculating

           statistics.
                                           37

-------
Chlorine is added to the water by the Town of Bernalillo prior to delivery to the storage tanks.
FTO personnel measured residual chlorine (total and free) and DO daily.  Table 4-16 shows the
residual chlorine and DO data for the feed and filtrate water.  During the verification test, the
total residual chlorine in the feed water ranged from 0.28 to 0.84 mg/L,  averaging 0.58 mg/L.
The feed water DO ranged from 5.53  to 8.57 mg/L, and averaged 7.01 mg/L. The filtrate water
averaged 0.50 mg/L total residual chlorine and 6.82 mg/L DO.  The free residual chlorine in the
feed water averaged 0.50 mg/L and the filtrate averaged 0.44 mg/L. The data from the 48-hour
intensive survey, presented in Table 4-17, is similar to the verification test results.
Table 4-16. Total and Free Residual Chlorine and DO
Date
4/18/06
4/19/06
4/20/06
4/21/06
4/22/06
4/23/06
4/24/06
4/25/06
4/26/06
4/27/06
4/28/06
4/29/06
4/30/06
5/01/06
Number of
samples
Average
Maximum
Minimum
Std. Deviation
95% Conf.
Interval
Free Chlorine
(mg/L)
Feed Filtrate
0.71
0.50
0.48
0.69
0.51
0.19
.(2)
0.35
0.47
0.56
0.51
0.52
0.54
0.50
13
0.50
0.71
0.19
0.13
(0.41-0.
0.84
0.43
0.41
0.59
0.37
0.42
(2)
0.16
0.39
0.40
0.44
0.42
0.45
0.41
13
0.44
0.84
0.16
0.15
60) (0.33-0.55)
Total Chlorine
(mg/L)
Feed Filtrate
0.84
0.60
0.54
0.74
0.57
0.28
.(2)
0.45
0.55
0.63
0.57
0.55
0.61
0.57
13
0.58
0.84
0.28
0.13
(0.48-0.67)
0.92
0.52
0.45
0.65
0.48
0.47
.(2)
0.23
0.45
0.47
0.51
0.45
0.50
0.46
13
0.50
0.92
0.23
0.15
(0.40-0.61)
DO
(mg/L)
Feed Filtrate
7.4
6.48
5.65
7.77
7.24
5.53
6.66
6.95
7.75
6.50
6.25
8.57
8.10
7.27
14
7.01
8.57
5.53
0.89
(6.41-7.
J1)
6.51
5.97
7.51
6.95
5.91
7.52
6.74
7.29
6.63
6.42
7.04
6.86
7.27
13
6.82
7.52
5.91
0.52
61) (6.44-7.19)
         collected.
   (2)No data collected due to equipment malfunction.
                                            38

-------
Table 4-17. Total and Free Residual Chlorine and DO Results for 48-Hour Survey


Free Chlorine
Time
Date (hours)
4/25/06
4/25/06
4/25/06
4/25/06
4/25/06
4/26/06
4/26/06
4/26/06
4/26/06
4/27/06
4/27/06
Number of samples
Average
Maximum
Minimum
Std. Deviation
95% Conf. Interval
0
1
3
6
12
18
24
30
36
42
48






(mg/L)
Feed Filtrate
0.35
0.54
0.5
0.55
0.54
0.57
0.47
0.53
0.53
0.56
0.56
11
0.52
0.57
0.35
0.06
(0.47-0.
0.16
0.41
0.41
0.46
0.45
0.44
0.39
0.39
0.38
0.45
0.4
11
0.39
0.46
0.16
0.08
57) (0.33-0.46)
Total Residual Chlorine

(mg/L)
Feed Filtrate
0.45
0.60
0.57
0.60
0.59
0.53
0.55
0.57
0.55
0.52
0.63
11
0.56
0.63
0.45
0.05
(0.52-0.
0.23
0.47
0.46
0.50
0.48
0.52
0.45
0.47
0.52
0.56
0.47
11
0.47
0.56
0.23
0.08
60) (0.40-0.53)


DO
(mg/L)
Feed Filtrate
6.95
-
-
-
-
-
7.75
-
-
-
6.50
3
7.07
7.75
6.50
NC
NC
6.74
-
-
-
-
-
7.29
-
-
-
6.63
3
6.89
7.29
6.63
NC
NC
  NC = Not calculated.

The results for the other water quality parameters are shown in Table 4-18. Statistical results are
presented in Appendix C.  The feed water concentrations were stable throughout the test. The
feed and filtrate water showed similar average concentrations of chloride, sulfate, TOC, fluoride,
calcium, magnesium, hardness, manganese,  iron,  and true color.  The ARS CFU-50 APC had
little or no impact on these water quality parameters.

Temperature was monitored daily in the feed and filtrate water.  The feed water averaged a
temperature of 22.4°C and the filtrate averaged 22.7°C.
                                          39

-------
Table 4-18. Other Water Quality Parameters
Parameter
Chloride
Sulfate
TOC
Fluoride
Calcium
Magnesium
Hardness
Manganese
Iron(1)
True Color
Units
mg/L
mg/L
mg/L
mg/L
mg/L
mg/L
mg/L
Mg/L
mg/L
C.U.
4/18/06
Feed Filtrate
180
110
0.3
0.3
74
11
230
<1
<0.02
-
180
110
0.3
0.3
74
11
230
<1
<0.02
-
4/19/06
Feed Filtrate
-
-
-
-
93 90
-
-
-
<0.02 <0.02
0 0
4/26/06
Feed Filtrate
180
110
0.2
0.3
93
12
282
<1
<0.02
5
170
100
<0.1
0.3
89
11
268
<1
<0.02
6
4/27/06
Feed Filtrate
170
110
<0.1
0.3
93
11
278
<1
<0.02
0
170
110
<0.1
0.3
84
12
259
<1
<0.02
1
 (; Iron was collected on a daily basis; all but one result were non-detect.

4.6     Operations and Maintenance Findings

FTO personnel operated the ARS CFU-50 APC during the 14-day verification period. FTO
personnel found it was easy to operate and required little time for daily maintenance. The FTO
was on site for two to three hours per day, with most of the time being spent performing ETV-
related activities, including flow checks, calibrations, and  similar activities. Typical monitoring
or maintenance on the unit would be rather minimal, and would primarily be related to cleaning
the filter press or performing routine inspections.

ARS provides an O&M manual for each  system. The draft O&M manual for the ARS CFU-50
APC, included in Appendix A,  provides a good description of the system, appropriate  safety
precautions, detailed descriptions  of operating procedures, the capability and operation of the
computer control system, and specific instructions for utility operators.

The  O&M manual provides detailed information on the various modes  that can be used for
operating the equipment. The modes are  preprogrammed operating conditions that include filter
backwash triggers and the manner in which the PLC responds to various signals and alarms. The
PLC discussion is thorough, and the programming provides good  operating flexibility for the
operator.

The  O&M manual also  describes the tanks, piping, and filter units, with information  on the
connections for each vessel.  Instructions for items to check prior to start-up are included in the
descriptions.

The  system is automated,  and  all equipment appeared  sturdy and properly selected for the
process.   Overall, the ARS  CFU-50 APC appears  well suited  to  small  or  medium-sized
installations where an operator is not present at all times, provided the filter press is sufficiently
sized and maintained.
                                          40

-------
The ARS CFU-50 APC includes a flow totalizer and flow rate meter for the filtrate water. The
system has  pressure gauges  on the feed and filtrate  lines that  provide  pressure data for
monitoring pressure differential (head loss) across  the filters.  This information is monitored by
the PLC and is available to the operator for review.

Findings  on specific components within the ARS CFU-50  APC are noted in the following
sections.

4.6.1   Electrical Consumption

The electrical use by the ARS CFU-50 APC is primarily for the reaction vessel, floe water pump
(downstream of the reaction vessel),  the clean water pump, waste pump, with negligible power
being consumed by a backwash pump, PLC, air compressor, and other instrumentation.  The test
system used  a 480 volts alternating current (VAC),  3 Phase, 60-ampere  and a 120 VAC,  1-phase,
20-ampere electrical supply. The test system had two 3 horsepower (hp) centrifugal pumps, one
to pump water from the reaction vessel which ran constantly, and one to pump clean water from
the device which ran  about 66% of the time  the device is  operational.  The reaction vessel
consumed power at an average of 4.2 volts and 30 amperes.  The waste pump is a 0.5 hp, 110
VAC pump,  which ran about 25% of the time the device, is operational.  The backwash pump is
a 5 hp pump, which ran less than 5%  of the time the device is operational.  The PLC and air
compressor  power  consumption was considered negligible.   Unadjusted  horsepower (not
adjusted for  efficiency factor) is equal to 746 watts per hp.  The itemized power consumption
usage, approximately 4.2 Kilowatt-hour (KwH), is outlined in Table 4-20.

Table 4-19.  Power Consumption
                       „         ,                  Power Consumption
                       Component                         „, TT, r
         	_	(KwH)	
          Reaction vessel                                  0.13
          Filter influent pump                              2.3
          Clean water pump                                 1.5
          B ackwash pump                                 0.19
          Waste pump                                     0.09
          PLC, air compressor                              Negligible
          Total power consumption	4.2	

4.6.2   Sand Filters

The ARS CFU-50 APC  is equipped with two sand filters, so one filter can operate while the
other  is in backwash  mode  or standby.  Backwash water  is stored in the backwash  water
reservoir, then filtered through a filter press, with the treated water returned to the reaction vessel
for re-treatment.  During the testing at this installation, there were no conditions  where the
pressure differential across both sand filters required that both filters backwash at the  same time.
One filter was always available for filtering the flocculent from the treated water.

Issues regarding the efficacy of the filtration process were noted during the verification test, as
shown in the turbidity data, with the filtrate water being consistently higher than the feed water.
                                          41

-------
As noted in Table 4-10 and Figure 4-3, on average, the filtrate turbidity was 0.8 NTU while the
feed turbidity was 0.3 NTU. Also, as summarized in Table 4-14 and shown in Figure 4-5, the
aluminum concentrations in the filtrate water were consistently higher than the concentrations in
the  feed water, with  the  aluminum  concentrations  in the filtrate water  exceeding the EPA
National Secondary Drinking Water Regulations. The turbidity and aluminum data indicate that
filtration mechanisms beyond  those  currently utilized in the ARS  CFU-50 APC would be
required to bring these concentrations closer to the feed water concentrations or within the EPA
secondary regulations.

4.6.3  Filter Press

Backwash waste is treated by a filter press designed  to remove the solids (flocculent) from the
backwash water. The filtrate from the filter press was transferred back to the reaction vessel for
re-treatment.  During the testing, when the flocculent caked in the filter press to a point where
water would no longer pass through it, the PLC shut down the entire system, as it is programmed
to do.  The procedures outlined in the O&M  manual provide clear instructions on cleaning the
filter press and resuming operation after cleaning. Page 65 of the O&M manual notes, "The use
and operation of the filter  press should not interfere with the normal and continuous use of the
treatment unit.. .However, the treatment unit will only run for a limited time with the filter press
disabled.  The filter  press is a vital  aspect of the treatment  unit  and must  be supported
appropriately."   Verification testing  substantiated the importance of the filter  press and its
appropriate maintenance as a critical aspect of the function of the ARS CFU-50 APC device.

4.6.4  Backwash Water Frequency  and Quality

The  ARS CFU-50 APC operates with an automated backwash sequence where backwash water
is passed through a filter press.  The backwash sequence uses water from the clean water storage
tank. The backwash water is stored in the flocculent water reservoir tank prior to treatment in
the filter press, where solids are removed and the filtered water is returned to the reaction vessel.
The  backwash cycle was  set for a fixed time duration of 120 seconds for backwash and 30
seconds for rinsing.  The combined backwash and rinsing resulted in approximately 250 gallons
of waste per backwash sequence.  Solids retained in the filter press are removed manually during
filter press maintenance. Since the ARS CFU-50 APC does not have a separate backwash waste
stream,  the evaluation of the backwash water frequency and quality during the verification was
limited  to TSS  analysis.   The  sample was collected from the flocculent water reservoir tank,
which was equipped with a mixer and a sample port near the bottom of the tank.  This sample
yielded  a concentration of 1,200 mg/L.

After the two weeks of operation, approximately 150 gallons of filter press sludge consisting of
hydrated floe was removed from the filter press.  The density of the floe is approximate 9.5
pounds  per gallon, so approximately  1,425 pounds of hydrated floe was created.  During this
timeframe, approximately 572,550 gallons of water was treated. By dividing the weight of the
hydrated flocculent by the volume of water treated, an approximate suspended solids (flocculent)
concentration of 300 mg/L  was generated by the reaction vessel.
                                          42

-------
Local disposal requirements determine whether filter press sludge is characteristically hazardous,
due to elevated arsenic (or other constituent) concentrations. A sample of the solids accumulated
over the  14-day test, which were placed in a 55-gallon drum and stored near the device, was
collected and  analyzed  by the Toxicity  Characteristic Leaching Procedure (TCLP) and the
California Waste Extraction Test (CAWET). This sample represented a composite of all solids
generated during the verification test. The filter cake sludge was not considered a hazardous
waste based on  the arsenic concentrations, which were below the  5 mg/L limit under the
Resource Conservation and Recovery Act (RCRA). Table 4-19 presents the results of the TCLP
and CAWET  analyses.   The laboratory test report received from TriMatrix Laboratories  is
included in Appendix F.

Table 4-20. Backwash Solids - TCLP and CAWET Analyses
Parameter
Arsenic
Barium
Cadmium
Chromium
Copper
Lead
Mercury
Nickel
Selenium
Silver
Zinc
Units
mg/L
mg/L
mg/L
mg/L
mg/L
mg/L
mg/L
mg/L
mg/L
mg/L
mg/L
TCLP
0.18
1.5
<0.010
0.74
0.23
<0.050
O.0002
<0.010
<0.10
<0.010
0.94
CAWET
3.7
<3.5
<0.01
2.5
3.1
<0.50
<0.0002
<0.10
<1.0
<0.10
<2.5
4.6.5   Programmable Logic Controller

The system PLC is designed to operate and monitor many of the operating functions  of the
device.  The PLC was not programmed to record data,  so readouts on component performance,
such as flow, pressure,  electrical settings, and other operational conditions had to be monitored
and recorded manually.  The PLC readings were easy to use, but required an understanding of
the PLC operating keys to display the readings.

The PSTP specified an evaluation of the length of filter runs, backwash cycle frequency, and the
pressure differentials across the  sand filters as part of the study. The PLC was not programmed
to record data on pressure differentials or the  time at which  backwash cycles were initiated.
Generally, this data would serve as the basis for such an evaluation. Without the PLC recording
this data, it would have had to be collected by the FTO technician manually monitoring the sand
filters, and recording the backwash start time and operating pressure. NSF concluded that, since
the ARS CFU-50 APC is designed  to remove  the flocculent material from the backwash and
return  the treated water back to the reaction  vessel  for treatment with little or no  operator
involvement, collection of the data  would  be  imprecise and difficult  to  evaluate, since
backwashing could have occurred during periods when FTO  personnel were not on site.
                                          43

-------
The PLC is designed to shut the entire system down in the event any sensor records a condition
that is outside preset  operating limits.  This condition was experienced four times during the
verification, resulting from the filter press being clogged to a point where no water passed
through it.  This resulted in high water levels in the Flocculent Water Reservoir, during which
the PLC shut the system  down as it was programmed to do.  During each shutdown condition,
after the filter press was cleaned, the alarm conditions in the PLC were cleared and the system
was restarted with no difficulty.

4.7    Quality Assurance/Quality Control

As described  in the PSTP, included in Appendix  B,   a  structured QA/QC  program was
implemented as part of this verification to ensure the  quality of the data being collected.  A
QAPP was developed as part of the PSTP and was followed by the field staff and laboratory
during the testing period.   Adherence to the established procedures ensured  that the data
presented in this report are sound, defensible, and representative of the equipment performance.

4.7.1   Documentation

The FTO recorded on-site data and calculations in a field logbook and prepared field log sheets.
Daily measurements were recorded on specially prepared data log sheets. The operating logbook
included calibration records for the  field  equipment used for on-site analyses. Copies of the
logbook, the daily data log sheets, and calibration log sheets are included Appendix D.

Data from the on-site laboratory and data log sheets were entered into Excel spreadsheets, which
were used to calculate various statistics (average, mean, standard deviation, etc.).  The data in the
spreadsheets were proofread by the initial data entry person and confirmed by NSF DWS Center
staff by a 100%  check  of the  data entries  to confirm  the information was  correct.  The
spreadsheets are presented in Appendix C.

Samples collected and delivered to the NSF Chemistry Laboratory for analysis  were tracked.
Each sample was assigned a location name, date, and time of collection, and the parameters were
written on the label.  The laboratory reported the analytical results using the NSF Chemistry
Laboratory management system reports.  These reports were received and  reviewed by the NSF
DWS Center coordinator. These laboratory data were entered by DWS Center personnel into the
data spreadsheets, corrected, and verified in the same manner as the field data. NSF laboratory
reports are included in Appendix E.

4.7.2   Quality Audits

The NSF QA department performed an on-site  audit on April 23 to review the field procedures,
including the collection of operating data and  performance of on-site analytical methods.  The
PSTP requirements and QAPP were used as  the basis for  the audit.  The NSF  QA  auditor
prepared an audit report, noting the following deficiencies:

   1.  The operations log does not indicate visitors to the site.
   2.  Photographs were not logged in the field logbook.
                                           44

-------
   3.  The standard concentrations used for the turbidimeter are not as listed on the checklist.
   4.  Pressure  gauge accuracy was checked by another calibrated pressure gauge  but not
       checked with a dead weight pressure tester as indicated in the checklist.
   5.  The checklist indicated the thermometer would be calibrated monthly.  However, the
       thermometer tag indicated it would be calibrated every six months.
   6.  The checklist indicates that the DO meter would be air calibrated.  However,  a liquid
       calibration method is used.
   7.  There are no Material Safety Data Sheets (MSDS) sheets at the site or the laboratory.
   8.  There is no eyewash station at the laboratory.

FTO personnel addressed deficiencies 1  and 2 immediately. Deficiencies 3 through 6 pertained
to correlating the calibration methods used by FTO personnel  with the documentation noted in
the PSTP and checklist. MSDS information for the laboratory was attainable  through NSF's on-
line database,  which was accessible to the FTO personnel via computer.  A portable eyewash
station was in the process of being procured for the laboratory, but the project concluded prior to
delivery.

The NSF QA Department reviewed the Chemistry Laboratory analytical results for adherence to
the QA requirements for calibration,  precision, and accuracy detailed in the  project QAPP, and
for compliance with the laboratory quality assurance requirements.   The laboratory raw data
records (run  logs,  bench sheets, calibrations records,  etc.)  are  maintained at NSF  and are
available for review.

4.7.3   Data Quality Indicators

The data quality  indictors established for the ETV project and described in the QAPP included:

   •   Representativeness;
   •   Accuracy;
   •   Precision; and
   •   Completeness.

4.7.3.1 Representativeness

Representativeness  refers to the degree to which the data accurately and precisely represent the
conditions or characteristics of the parameter represented by the data.  In this verification testing,
representativeness  was ensured by  FTO  personnel  executing  consistent  sample  collection
procedures in  accordance with established approved procedures, and following specific sample
preservation,  packaging,  and delivery procedures.  Approved  analytical methods were used to
provide results that represent the accurate and precise  measurements of drinking water.   For
equipment  operating data, representativeness entailed collecting and documenting a sufficient
quantity of data during operation to be  able to detect a change in operations. For most water
treatment processes involving total arsenic removal, detecting a ±10% change in an operating
parameter is sufficient.   The primary operating parameter for this verification test was filtrate
volume treated per day and water quality (e.g. total arsenic concentrations, fouling  parameter
concentrations, etc.). For this verification, the total arsenic concentrations were somewhat lower
                                           45

-------
than the concentrations measured during prior sampling  events  (see Table  1-1), but the
concentrations were consistent, and other parameters were within ranges that would not impact
the performance  of the ARS  CFU-50 APC  device.   Thus, these data were  judged to be
representative and were included in the data set for the verification test.

4.7.3.2 Accuracy

On-Site Equipment Accuracy and Calibration

On-site equipment,  including  ARS CFU-50 APC flow meters and DWTS on-site  analytical
equipment, were tested for accuracy through regular calibration checks. Meters and gauges were
checked at the frequencies presented in Table  4-20.  The  calibration records for pH, turbidity,
total and free residual chlorine, and DO were recorded in the field calibration log (Appendix D).
Calibrations were performed  at the frequency  required,  and were within the  specified QC
objectives on all days analyses were performed.

The ARS CFU-50 APC had  a filtrate water  flow  rate and totalizer meter. The "bucket and
stopwatch" technique was used to determine the accuracy of the flow meters. Table 4-21 shows
the calibration data.  All  calibrations were within the defined objective of ±10%.
Table 4-21. Field Instrument Calibration Schedule
Instrument
Pressure Gauges
Flow Meter
Totalizer Meter
Bench Top Turbidimeter
Calibration Method
dead weight calibration tester,
evaluation against another calibrated
gauge, or manufacturers certification
volumetric "bucket & stop watch"
volumetric "bucket & stop watch"
secondary turbidity standards
primary turbidity standards
Frequency
once during
testing
weekly
weekly
daily
weekly
Acceptable
Accuracy
± 10%
± 10%
± 1.5%
PE sample
 Portable pH/ISE Meter
 with Combination pH/
 Temperature Electrode

 Portable Colorimeter

 Thermometer (National Institute
 of Standards and Technology
 (NIST)-traceable)

 DO
three-point calibration using
4.0, 7.0 and 10.0 buffers
chlorine check standard

calibration against NIST traceable
air calibration method or zero
method, as recommended by the
meter manufacturer
 daily
 daily

 twice
annually
 daily
 ±5%



± 25%

 ±5%



± 10%
                                           46

-------
Table 4-22. Flow Meter Calibration Data
Date
4/21/06
4/27/06
4/30/06
Feed
Calibration Flowmeter
Result Reading
(gpm) (gpm)
37.3
36.4
36.6
36.3
35.1
34.9
Filter Influent
Calibration Flowmeter
Result Reading
(gpm) (gpm)
32.8
31.3
34.1
34.3
32.6
35.8
Laboratory Analyses

Accuracy for the laboratory analyses is quantified as the percent recovery of a parameter in a
sample to which  a known quantity  of that parameter was added.  Equation  4-1  is used to
calculate accuracy:

          Accuracy = Percent Recovery = 100 x [(Xkn0wn - Xmeasured) + X^own]    (4-1)

          where    Xkn0wn   = known concentration of measured parameter
                           = measured concentration of parameter
Accuracy also incorporates calibration procedures and use of certified standards to ensure the
calibration curves and references for analysis are near the "true value." Accuracy of analytical
readings is measured through the use of spiked samples and laboratory control samples (LCS).
The percent recovery is calculated as a measure of the accuracy.

The QAPP and the NSF Chemistry Laboratory QA/QC requirements established the frequency
of spike sample analyses at 10% of the samples analyzed.  LCS are also  run at a frequency of
10%.  The recovery  limits specified  for the parameters in this verification were 70-130% for
laboratory -fortified samples  and 85-115% for LCS.  The NSF QA department reviewed the
laboratory records and found all analyses for all sample groups were within the QC requirements
for recovery.  Calibration requirements were also achieved for all analyses.

The arsenic speciation resin columns were tested to ensure proper separation and recovery of the
arsenic species. Each lot of the arsenic  speciation resin was checked once against samples with
known concentrations of As (III) and As (V). This QC check assured that the resin was properly
prepared. The NSF Chemistry Laboratory maintained the documentation for the column checks.

4. 7.3.3 Precision

Precision refers to the degree of mutual agreement among individual measurements and provides
an estimate  of random error.  Analytical  precision  is a measure of how far an individual
measurement may be from the mean of replicate  measurements.  The relative standard deviation
recorded from sample analyses was  used to quantify  sample precision.  The percent relative
standard deviation was calculated using the equation presented as Equation 4-2:
                                          47

-------
             Percent Relative Standard Deviation = S(100) / Xaverage     (4-2)

             where: S      = standard deviation
                    Xaverage = the arithmetic mean of the recovery values
Standard Deviation is calculated in Equation 4-3:
Standard Deviation =                 I ±>   (Xi-X)2                (4-3)
             where: X;     = individual measured values
                    X     = arithmetic mean of the measured values
                    n      = number of determinations

Acceptable analytical  precision for the verification test was set at a percent relative standard
deviation  (%RSD) for drinking water  samples of 30%.   Field  duplicates were collected to
incorporate both sampling  and analytical  variation  to  measure overall precision against this
objective.  The laboratory precision for the methods selected was tighter than the 30% overall
requirement, generally set at 20% based on the standard NSF Chemistry  Laboratory method
performance.

Field Duplicates

Field duplicates were  collected for all analyses (field lab and analytical laboratory)  to monitor
overall precision.  The field duplicates included samples for both  sample  locations:  feed and
filtrate water.

Tables 4-21  and 4-22  summarize the results for the field duplicate samples.  The precision for
analyses performed in the laboratory, as measured by these field duplicates, met the overall QC
objective of 30% RSD for  most samples.   All precision values for the arsenic duplicate data,
except for one arsenic III set with a %RSD of 85%, met the QC objective of 30% RSD.  Three
aluminum and both TOC duplicates were above the maximum precision of  30%. It is unknown
why the three  aluminum samples had high precision (54-110%); however,  the verification test
data also had high  variation of aluminum in the feed water, ranging from <10 |ig/L to 84 |ig/L.
The high precision values for the TOC samples can be attributed to the low concentrations of
TOC.

The field  analyses data  for field duplicates were acceptable for all parameters. The true  color
data had two precision values of 141%, however, this was due to one sample reading at 0 C.U.
and the other sample at 1 C.U.
                                           48

-------
Table 4-23. Precision Data - Field Duplicates for Laboratory Parameters

                                                Total Arsenic
Date
4/18/06
4/24/06
5/01/06
Rep 1
14
11
12
Feed Water
Rep 2
15
13
12
%RSD
4.9
12
0
Rep 1
6
6
6
Filtrate
Rep 2
5
6
7
%RSD
13
0
11
                                              Dissolved Arsenic
Date
4/18/06
5/01/06
Rep 1
12
11
Feed Water
Rep 2
12
11
%RSD
0
0
Rep 1
18000(1)
4
Filtrate
Rep 2
4
5
%RSD
16
                                                  Arsenic III
Date
4/18/06
5/01/06
Rep 1
20
12
Feed Water
Rep 2
5
10
%RSD
85
13
Rep 1
8
4
Filtrate
Rep 2
7
5
%RSD
9
16
                                             Alkalinity (mg/L as CaCO3)
Date
4/18/06
5/01/06
Rep 1
130
120
Feed Water
Rep 2
140
150
%RSD
5
16
Rep 1
Filtrate
Rep 2
%RSD
- -
                                                  Aluminum
Date
4/18/06
4/24/06
5/01/06
Rep 1
71
79
13
Feed Water
Rep 2
75
<10
<10
%RSD
3.9
110
18
Rep 1
540
780
890
Filtrate
Rep 2
240
650
240
%RSD
54
13
81
                                                  Other Parameters
Parameter
Chloride (mg/L)
Sulfate (mg/L)
Calcium (mg/L)
Magnesium
(mg/L)
Manganese
Fluoride (mg/L)
Iron (mg/L)(2)
TOC (mg/L)
Rep 1
180
110
74
11

<1
0.3
<0.02
0.3
Feed Water
Rep 2
180
120
76
10

<1
0.3
<0.02
0.1
%RSD
0
6
1.9
6.7

0
0
0
71
Rep 1
180
110
74
11

<1
0.3
<0.02
0.3
Filtrate
Rep 2
180
110
76
10

<1
0.3
<0.02
<0.1
%RSD
0
0
1.9
6.7

0
0
0
71
Note: For the statistical calculations, all non-detect data were used as the minimum reporting limit.
(1)  This dissolved arsenic concentration was considered an outlier and was not used for statistical purposes.
(2)  Additional iron duplicate samples were collected; all duplicate set results were non-detect.
                                                49

-------
Table 4-24. Precision Data - Field Duplicates for Field Parameters
                                                pH (S.U.)
Date
4/18/06(1)
4/25/06
4/26/06
4/30/06
Rep 1
7-8
7.7
7.7
7.7
Feed Water
Rep 2
7-8
7.8
7.8
7.7
%RSD
0
0.91
0.91
0
Rep 1
7.0
7.7
7.7
7.9
Filtrate
Rep 2
7.0
7.6
7.8
7.9
%RSD
0
0.92
0.91
0
                                            Temperature (°C)
Date
4/18/06
4/25/06
4/30/06
Rep 1
20.2
22.8
23.2
Feed Water
Rep 2
20.2
22.8
23.2
%RSD
0
0
0
Rep 1
20.8
23.5
23.2
Filtrate
Rep 2
20.8
23.5
23.2
%RSD
0
0
0
                                            True Color (C.U.)
Date
4/19/06
Rep 1
0.0
Feed Water
Rep 2
1
%RSD
141
Rep 1
0.0
Filtrate
Rep 2
1
%RSD
141
                                       Turbidity (Bench Top) (NTU)
Date
4/18/06
4/30/06
Rep 1
0.47
0.18
Feed Water
Rep 2
0.51
0.15
%RSD
5.8
13
Rep 1
0.79
0.37
Filtrate
Rep 2
0.83
0.35
%RSD
3.5
3.9
                                       Free Residual Chlorine (mg/L)
Date
4/18/06
4/25/06
4/30/06
Rep 1
0.71
0.35
0.54
Feed Water
Rep 2
0.71
0.35
0.42
%RSD
0
0
18
Rep 1
0.84
0.16
0.45
Filtrate
Rep 2
0.84
0.16
0.50
%RSD
0
0
7
                                      Total Residual Chlorine (mg/L)
Date
4/18/06
4/25/06
4/30/06
Rep 1
0.84
0.45
0.61
Feed Water
Rep 2
0.84
0.45
0.61
%RSD
0
0
0
Rep 1
0.92
0.23
0.50
Filtrate
Rep 2
0.92
0.23
0.52
%RSD
0
0
3
                                               DO (mg/L)
Date
4/18/06
4/30/06
Rep 1
7.40
8.1
Feed Water
Rep 2
7.42
7.74
%RSD
0.19
3.21
Rep 1
6.50
6.86
Filtrate
Rep 2
6.55
6.46
%RSD
0.54
4.25
      pH samples analyzed with Litmus paper due to instrument malfunction.
                                           50

-------
Laboratory Analytical Duplicates

The  NSF Chemistry Laboratory precision was monitored  during  the verifications test  in
accordance with QAPP and the NSF quality assurance program.  Laboratory duplicates were
analyzed at  10% frequency of samples  analyzed.   The NSF QA  department  reviewed the
precision information and determined that the laboratory data met QC precision requirements.

4.7.3.4 Method Blanks

The  laboratory included method blanks as part of the standard  analysis procedures.  Method
blanks were analyzed in accordance with the  approved methods.  The NSF QA department
reviewed the laboratory data  and found the  method blanks to be acceptable.   No data were
flagged as having been affected by method blank results.

4.7.3.5 Completeness

Completeness is defined as the following (Equation 4-4) for all measurements:

                       %C = (V/T)X100                        (4-4)

      where:  %C    = percent completeness
              V     = number of measurements judged valid
              T     = total number of measurements

Completeness  refers to the amount of valid, acceptable  data collected from  a measurement
process compared to the expected amount to be obtained.

The  completeness objective for data generated during this verification test was based on the
number of samples collected and analyzed for each parameter and/or method. A completeness
objective of 90% applied to: arsenic, aluminum, alkalinity, iron, temperature, pH, daily bench
top turbidity, residual chlorine, and  DO.  Samples  for these parameters were  collected and
analyzed at the frequency specified in the PSTP and QAPP for the verification test.  All of the
weekly  parameters  met or exceeded the completeness objective of 80%. A  completeness
objective of 90% applied to the following operating parameters: feed and filtrate flow rate and
pressure differential across the filter.  The completeness objective was met for these parameters.
Table 4-24 provides a summary of the completeness results for the verification test.
                                          51

-------
 Table 4-25. Completeness Results
        n     ±               Percent                         „       ,
        Parameter          „    .                            Comment
	Completeness	
  .    .                           1 „ „       All scheduled samples and analyses completed for total
                                           and speciation requirements.
 Aluminum                       100       All scheduled samples and analyses completed.
 Iron                            100       All scheduled samples and analyses completed.
 pH                             100       All required daily measurements recorded.
 Bench top Turbidity               100       All required daily measurements recorded.
 „,_.,,„,,.              __        One daily analysis missed due to equipment
 Total Residual Chlorine            93           ,,,   /
                                           malfunction.
 T-,   T>   -j  i m.  •               n-3        One daily analysis missed due to equipment
 Free Residual Chlorine            93           ,,,   /
                                           malfunction.
 T-,   ,   j TTU  *  el   n *       m mn      All required daily measurements recorded, except for
 Feed and Filtrate Flow Rate      93-100          r-u  ^  n     .     A  -i o« ™n^
                                           one nitrate flow rate on April 25, 2006.
 Feed and Filtrate Pressure	100	All required daily measurements recorded.	

 4.7.4   Effect of Sample Preservative on Arsenic Speciation

 The arsenic speciation  data  in the  feed water reports  variable concentrations  of arsenic III
 ranging from non-detectable to 20 |ig/L.  This data appeared anomalous, given the consistent total
 arsenic concentrations in the feed water  ranging from 11  to 14 |ig/L. The arsenic speciation was
 conducted in the field using an acceptable method by the EPA and was audited by NSF QA. The
 field speciation method  requires filtration of the sample  and preservation with nitric or sulfuric
 acid.

 The feed water was  chlorinated (0.7 mg/L free chlorine) and had a high pH (7.5 - 7.8  S. U.).
 The feed water  had no detectable  concentrations of  iron.   Aluminum  was  detected  at  a
 concentration of  27  mg/L. The likely source of the aluminum was the ARS reaction vessel or
 filter press. Chlorination probably facilitated the rapid oxidation of arsenic (III) to (V) such that
 even short time delays in field speciation could result in variable speciation data exhibited in the
 report.  In the speciation method, the sample  is filtered  and preserved with sulfuric acid to fix
 arsenic (III).  With rapid oxidation induced by chlorination, the time to filter the  sample could
 affect speciation results.

 In  order to verify the  arsenic speciation  results,  NSF conducted a brief study to  ascertain
 whether the time of preservation affected arsenic  speciation  concentrations.   Samples were
 collected in the field in which one sample was unpreserved and the other  sample was preserved
 with sulfuric acid to fix arsenic (III).  Samples were then sent to NSF's laboratory.

 Upon arrival  at the laboratory, speciation  was performed  on the  samples.  The samples were
 filtered and passed though the resin column.   The results indicated that the sample preserved
                                             52

-------
immediately in the field with sulfuric acid had arsenic (III) at a concentration of 15 ng/L, while
the unpreserved water had a non-detectable arsenic (III) concentration.

This evaluation suggests that the arsenic (III) data are related to the rate of oxidation due to the
chlorination, and the inherent  variation of chlorine dosing time  versus  sample collection and
speciation.  Even a  small  amount of delay in field  speciation steps may have influenced the
arsenic speciation result.

4.7.5  Deviations from PSTP

During  testing, the  FTO implemented some testing procedures  or methodologies that were
different from those specified in the PSTP.  These deviations are addressed in various sections
throughout the verification report, and are summarized in Table 4-26, and do not appear to have
any impact on the findings or conclusions in this verification.

Table 4-26. Deviations from PSTP
 PSTP
Section
  No.
        Description
                 Reason for Deviation
  3.5
 4.4.3
  4.5
4.5.5.4
Evaluation of length of filter
runs between backwash cycles
and change in pressure across
filter media over time

The arsenic loss test was to be
conducted prior to starting the
verification test (Test C).
Test C will be run for a
continuous 320-hour period.
Evaluation of suspended solids
in backwash limited to one
sample.	
The system PLC was not designed to record data on when the
backwash cycles are initiated.  Manual recording of pressure
would have provided imprecise information of filter run time
and pressure increases overtime.

ARS personnel operated the system prior to the ETV
verification, and indicated that the system was ready for
operation once the FTO personnel mobilized to the test site.
NSF, ARS, and the FTO agreed that the operating conditions
configured by ARS were sufficient to begin the verification
test (Test C) without first conducting the arsenic loss test.

The filter press caused conditions where the system would
shut down  resulting in only 287 hours during the 14-day
period of evaluation.

The ARS CFU-50 APC does not have a separate discharge
mechanism for backwash. Solids were evaluated by
weighing the solids extracted from the filter press.	
                                            53

-------
                                    Chapter 5
                                    References

EPA/NSF ETV Protocol for Equipment Verification Testing for Arsenic Removal, U.S. EPA/NSF
International. September 2003.

Product Specific Test Plan for the Advanced Remediation Systems USA, LLC ARS CFU-50 APC
Electroflocculation and Filtration Water Treatment System for Arsenic Removal from Drinking
Water. NSF International. March 2006.

Standard Methods for the Examination of Water and Wastewater, 20th edition, APHA, AWWA,
and WEF, Washington D.C. 1999.
                                         54

-------