Standard Operating Procedure for
Analysis of Sediment for Total Mercury Using
  the Cold Vapor Technique with the Leeman
        Labs, Inc. Automated Mercury System
             Theresa Uscinowicz, A & O Chemical Company1
                                               and
                             Ronald Rossmann, USEPA
                           Large Lakes Research Station
                                     9311 Groh Road
                                  Grosselle, MI48138

                   Mid-Continent Ecology Division - Duluth
 National Health and Environmental Effects Research Laboratory

                                     Octobers, 1996

                                          Revision 1
    1 Current affiliation is SoBran, Inc./Pathology Associates International.

-------

-------
    Standard Operating Procedure for Analysis of Sediment for Total
  Mercury Using the Cold Vapor Technique with the Leeman Labs, Inc.
                         Automated Mercury System


1.0    Introduction

       Elemental concentrations of mercury in sediment and water are determined by the PS200 system,
       and its operation  is based upon cold vapor AAAS. The prepared sample enters the system in the
       divalent phase, and is mixed with stannous chloride to form elemental mercury vapor.  This
       mixture moves to the liquid gas separator, and argon carries the mercury vapor through a drying
       tube for vapor removal.  The vapor enters one path of the cell optimized. The mercury lamp emits
       light at 254 nm, and absorbance is measured by the detector.

2.0    Materials Required

2.1     Chemicals

       Reagents needed include the following ultra pure grade chemicals:

       2.1.1   Leeman Labs 1 OOppm Mercury Standard

       2.1.2   Leeman Labs Hydrochloric, Nitric Acids

       2.1.3   Leeman Labs Ultra Pure Water

       2.1.4   Liquinox

       2.1.5   J.T. Baker Brand Hydrochloric, Nitric Acids

       2.1.6   J.T. Baker Brand Stannous Chloride or Leeman Labs Stannous Chloride

       2.1.7   Hydroxylamine Hydrochloride

       2.1.8   Magnesium Perchlorate 10-20 size mesh

       2.1.9   Potassium Permanganate

2.2     Equipment and Supplies

       2.2.1   Supplies Needed for analyses  include:

             2.2.1.1              PS200 Automated Mercury Analyzer with autosampler,
                                with


                                       2-475

-------
SOP for Analysis of Sediment for Total Mercury
Using the Cold Vapor Technique with the
Leeman Labs, Inc. Automated Mercury System	Volume 2, Chapter 2

                                   associated data acquisition system
              2.2.1.2              AP200 Automated Preparation System with associated data
                                                   acquisition system


              2.2.1.3              Analytical Balance in Biology Lab, Mettler 2100T


              2.2.1.4              EDP pipettors and associated disposable tips


              2.2.1.5              Associated pump tubing for sample drainage, tin chloride


              2.2.1.6              Polyethylene tubes 12 mL capacity, and caps


              2.2.1.7              Polyethylene tubes 45mL capacity


              2.2.1.8              Teflon or polyethylene beakers


              2.2.1.9              Teflon wash bottles


              2.2.1.10       Teflon bottles (60 mL)


              2.2.1.11       Electronic balance


              2.2.1.12       PVC gloves


              2.2.1.13       Paper towels, clean wipes


              2.2.1.14       Lubricating oil for autosampler


              2.2.1.15       Quartz wool and quartz glass drying tubes


              2.2.1.16       Teflon spatula


       2.2.2   Supplies needed before sample analyses if not using automated preparation system:

              2.2.2.1         CEM microwave digester with associated Teflon PFA vessels

              2.2.2.2         Low density 30-mL polyethylene bottles for sample storage

2.3    Reference Documents

       The user of this method must be familiar with the following established standard
       operation procedures:

       LLRS-MET-SOP-001         Standard Operating Procedures for the Preparation of Materials
                                   used for Ultra-low Trace Element Analyses
                                            2-476

-------
                                               SOP for Analysis of Sediment for Total Mercury
                                                    Using the Cold Vapor Technique with the
 Volume 2, Chapter 2                                     Leeman Labs, Inc. Automated Mercury
	System
                                        2-477

-------
SOP for Analysis of Sediment for Total Mercury
Using the Cold Vapor Technique with the
Leeman Labs, Inc. Automated Mercury System	Volume 2, Chapter 2

       LLRS-MET-SOP-003         Standard Operating Procedures for the Maintenance of the LLRS
                                  Trace Metal Laboratories

       LLRS-MET-SOP-010         Standard Operating Procedures for Analysis of Total Mercury in
                                  Tissue and Sediment using the Cold Vapor Technique with the
                                  Perkin-Elmer Model MHS-20 Gold Amalgam System

       LLRS-QA-001               Minimum Analytical Quality Assurance Objectives for U. S. EPA
                                  Large Lakes Research Station

       LLRS-QA-SOP-001          Standard Operating Procedures for the Release of Data

       LLRS-QA-SOP-002          Standard Operating Procedures for the Routine Review of Data
                                  Quality and Quantity

3.0   Reagent Preparation

3.1     5% Nitric Acid Solution

       This acid solution is used for rinsing materials used for cleaning materials used in the
       analysis of samples.

       3.4.1   Rinse a pre-cleaned graduated cylinder (100 mL) with three rinses of MSQ.

       3.4.2   To the graduated cylinder, add 95 mL of MSQ.

       3.4.3   Carefully add 5 mL of concentrated reagent grade nitric  acid.

       3.4.4   Transfer the solution to a squirt bottle.

       3.4.5   Repeat steps 3.4.1  through 3.4.3 but add five milliliters of reagent grade acid.  (For
              preparation of 5%  nitric acid rinses).

       3.4.6   Repeat step 3.4.4 but add to separate precleaned teflon bottle.

3.2     10% Hydrochloric Acid Rinse  Solution

       Use reagent grade J.T. Baker or Leeman Labs hydrochloric acid. The amount needed for
       a one day run is 300 mL.  Approximately 2 L will be needed to be prepared weekly.

                                           WARNING

       Hydrochloric acid is highly corrosive and incompatible with metals, hydroxides,
       amines, and alkalis.  Handle it while wearing personal protection gear.  A full face
       shield is recommended. Always use the concentrated acid under a fume hood.  Store it
                                          2-478

-------
                                                  SOP for Analysis of Sediment for Total Mercury
                                                       Using the Cold Vapor Technique with the
     Volume 2, Chapter 2                                   Leeman Labs, Inc. Automated Mercury
	System

       in an appropriate place.  Store concentrated acid in a corrosives cabinet.
       3.2.1   Use a precleaned 1000 mL polyethylene graduated cylinder.  Carefully rinse the
              inside of the cylinder with 5% reagent grade nitric acid using a Teflon (precleaned)
              squirt bottle. Follow this rinse with three rinses of Millipore Super-Q water
              (MSQ).

       3.2.2   Add 900 mL of MSQ to the graduated cylinder.

       3.2.3   Carefully add 100 mL of reagent grade hydrochloric acid to the cylinder.

       3.2.4   Transfer the mixture to a high density polyethylene bottle from Leeman
              Labs.

3.3    1:1 Nitric Acid Solution

       As  recommended by the manufacturer,  polyethylene autosampler cups must soak in 1:1
       nitric acid before running to ensure acceptable results. Soak the cups for at least two
       hours. Use J.T.  Baker trace metal grade nitric acid for preparation. This solution can be
       recycled, provided the autosampler cups are adequately rinsed with MSQ a minimum of
       ten times before and after use. Dispose  of 1:1 nitric after one month to eliminate the
       possibility of any residual contamination.

                                           WARNING

       Nitric acid is corrosive and incompatible with combustible materials, metallic powders,
       hydrogen sulftde, carbides, and alcohols.  Handle it with personal protection. A full
       face shield is recommended for the concentrated acid.  Always use the acid under a
       fume hood.  Store the concentrated acid in a corrosives cabinet.

       3.3.1   Use an appropriate pre-cleaned container.  An empty Suprapure hydrochloric  acid
              bottle has been used.  Rinse the container at least three times with MSQ before
              addition of nitric acid.

       3.3.2   Using the graduations on the glass bottle, add 400 mL MSQ.

       3.3.3   Carefully add 400mL of J.T. Baker nitric acid to the glass bottle.

       3.3.4   Swirl the contents of the bottle and , if necessary, label.

3.4    10%(w/v) Stannous Chloride Solution

       The volume of tin chloride solution required to run is dependent upon the daily run time.
       To run for six hours approximately 250  mL are required.
                                           2-479

-------
SOP for Analysis of Sediment for Total Mercury
Using the Cold Vapor Technique with the
Leeman Labs, Inc. Automated Mercury System	Volume 2, Chapter 2
                                            2-480

-------
                                                  SOP for Analysis of Sediment for Total Mercury
                                                        Using the Cold Vapor Technique with the
     Volume 2, Chapter 2                                    Leeman Labs, Inc. Automated Mercury
	System

                                        WARNING

       Stannous chloride should be handled with care.  Avoid contact with eyes, skin, and
       clothing.  Avoid breathing its dust. Handle solid chemical under a fume hood.
       Handle while wearing personal protection gear for eyes and skin.

       3.4.1   Rinse a precleaned wide mouth teflon bottle three times with quartz distilled water.

       3.4.2   Tare the bottle and add 25 g of tin chloride to the bottle.

       3.4.3   Rinse a precleaned graduated cylinder with 5% reagent grade nitric acid solution
              followed by three rinses of MSQ.

       3.4.4   Carefully add 25 mL of reagent grade hydrochloric acid to the graduated cylinder.
              Carefully pour the cylinder's contents into the bottle containing the stannous
              chloride.

       3.4.5   Swirl the contents of the bottle vigorously to dissolve the stannous
       chloride.

       3.4.6   After the stannous chloride has been dissolved, add 200 mL of MSQ using the
              same graduated cylinder.

       3.4.7   Replace the cover and vigorously shake the bottle to ensure all of the tin chloride
              is dissolved.

3.5    10% Nitric Acid Solution

       This acid solution is used for preparation of standards and for dilution of samples.
       Historically, this solution is used for samples that have undergone microwave digestion.

       3.5.1.  Rinse a pre-cleaned graduated cylinder (100 mL) with 5% reagent grade  nitric
              acid solution followed by three rinses of MSQ.

       3.5.2   To the graduated  cylinder, add 90 mL of MSQ.

       3.5.3   Carefully add 10 mL of concentrated Seastar or reagent grade nitric acid
              (whichever matches the matrix of samples).

       3.5.4   Transfer the solution to a teflon bottle.
                                           2-481

-------
SOP for Analysis of Sediment for Total Mercury
Using the Cold Vapor Technique with the
Leeman Labs, Inc. Automated Mercury System	Volume 2, Chapter 2

3.6    Working Mercury Standard Solution

       Commercial 100 ppm Leeman Labs mercury standard is used to prepare the working
       standard.  The working standard is prepared at a concentration of 100 ppb or 0.1 jig
       Hg/mL and is made fresh weekly.

                                       WARNING

       Mercury is a poison.  Handle it, its compounds, and its solutions with personal
      protection. Mercury can form a vapor and be inhaled. It also is absorbed through the
       skin.  Use this material in a fume hood. Always wear personal protection gear.

       3.6.1   Rinse a pre-cleaned teflon bottle (LLRS-MET-SOP-001) three times with MSQ
              and air dry.

       3.6.2   Using a precleaned graduated cylinder, rinse three times with 5% nitric
              and three times with MSQ.

       3.6.3   Using two EDP separate automatic pipettors, rinse 2-1000 mL tips three times
              with a 5% reagent grade nitric solution followed by three rinses with MSQ.

       3.6.4   Tare a dry bottle on the electronic scale.

       3.6.5   Using an additional EDP pipettor, rinse a 100 jiL tip in the same manner as 3.6.2

       3.6.6   Using the  cleaned  100 jiL tip, carefully add 50 jiL of the commercial 100 ppm
              Leeman Labs mercury standard solution to the bottle.

       3.6.7   Transfer 45mLs of the 10% nitric acid solution using the cleaned graduated
              cylinder.

       3.6.8   Using an EDP pipettor, with a 1000 jiL precleaned tip, add 10% nitric acid
              solution to the bottle until the weight is 50 g.

3.7    Preparation of Recommended Range of Calibration Standards

       The calibration standards are prepared fresh  semi weekly.

       3.7.1   Microwave Digestion Standard Range

              The following range of standards  has been used for analyses of Green Bay
              Sediment samples and bracket the samples well.  0.250 ppb, 0.500 ppb, 1.00 ppb
              2.00 ppb,  5.00 ppb are used (0.00025 |ig/mL, 0.00050 |ig/mL, 0.001  |ig/mL,
              0.002 |ig/mL, 0.005 jig/mL). It may be possible to go below 0.250 ppb depending
              upon instrument performance. The lowest concentration above background noise
                                          2-482

-------
                                              SOP for Analysis of Sediment for Total Mercury
                                                   Using the Cold Vapor Technique with the
 Volume 2, Chapter 2                                     Leeman Labs, Inc. Automated Mercury
	System

         is approximately 0.087 ppb (0.000087 jig/ mL). 60 mL of each calibration
         standard will last for two daily runs. Prepare  double of the desired standard that
         will be run as a check standard.  The autosampler cups must be filled to at least 40
         milliliters, (60 mL total capacity).

         3.7.1.1         Rinse six pre-cleaned teflon bottles three times with MSQ and
                         allow to air dry.

         3.7.1.2         Prepare two separate EDP pipettors each with a precleaned 1000
                         |iL tip. (Rinse each pipette tip three times with 5% nitric acid
                         followed by three rinses with MSQ) .

         3.7.1.3         Rinse a precleaned graduated cylinder with 5% nitric acid
                         followed by three rinses of MSQ.

         3.7.1.4         For each standard, tare each bottle individually.

         3.7.1.5         For the 0.250 ppb standard, pipette 150 pL of the 100 ppb
                         working standard to the bottle.

         3.7.1.6         Using the graduated cylinder, bring the total volume up to 55 mL
                         by the careful addition of prepared Seastar 10% nitric acid. The
                         weight of the bottles contents should  now be roughly 55 g.

         3.7.1.7         Add prepared Seastar 10% nitric acid with the other EDP
                         pipettor and precleaned tip until the total weight is 60 g.

         3.7.1.8         Repeat steps 3.7.1.6 - 3.7.1.9 for the remaining standards by the
                         addition of 300 \iL of the 100 ppb standard for a 0.500 ppb
                         calibration standard, 600 jiL of the 100 ppb standard for the 1.00
                         ppb calibration standard, 1200 jiL of the 100 ppb standard for the
                         2.00 ppb calibration standard, and 3000 jiL of the 100 ppb
                         standard for the 5.00 ppb calibration standard.

         3.7.1.9         Do not recycle the standards remaining in the autosampler cups at
                         the end of the day. Properly dispose of these daily. Attempts to
                         recycle the standards from autosampler cups have given
                         diminished intensities.

   3.7.2  Automated Digester Standard Range

         Use the following range of standards: 0.000 ppb, 0.125 ppb,  0.250 ppb, 0.500 ppb,
         1.00 ppb, 2.00 ppb. Prepare in a 2% hydrochloric acid matrix in precleaned Teflon
         bottles. Use Leeman Labs or J.T. Baker hydrochloric acid. Prepare 50mL of each
                                       2-483

-------
SOP for Analysis of Sediment for Total Mercury
Using the Cold Vapor Technique with the
Leeman Labs, Inc. Automated Mercury System	Volume 2, Chapter 2

              standard weekly or as needed.

              3.7.2.1          0 ppb Standard

                     3.7.2.1.1      Rinse a precleaned graduated cylinder with 5% nitric acid
                                   followed by three  rinses of MSQ.  Dispose of  waste in an
                                   appropriate container.

                     3.7.2.1.2      Rinse a precleaned SOOmL teflon bottle three times with
                                   MSQ.

                     3.7.2.1.3      Add 245 mL MSQ to the bottle.

                     3.7.2.1.4      Carefully add 5 mL of concentrated hydrochloric acid to the
                                   bottle.

                     3.7.2.1.5      Transferthe solution to a teflon bottle.

              3.7.2.2          0.125 ppb Standard

                     3.7.2.2.1      Prepare one EDP pipettor with a precleaned 100 jiL
                                   tip.

                     3.7.2.2.2      Prepare two separate EDP pipettors each with a precleaned
                                   1000 pLtip.

                     3.7.2.2.3      Rinse a teflon bottle three times with MSQ that will be used
                                   for each standard.

                     3.7.2.2.4      Tare the bottle on the balance.

                     3.7.2.2.5      Pipette 62.5 jiL of the  100 ppb working standard into the
                                   bottle

                     3.7.2.2.6      Use the rinsed graduated cylinder to transfer 45 mL of the 0
                                   ppb standard to the bottle.

                     3.7.2.2.7      Slowly pipette 5 mL of 0 ppb standard to the bottle until the
                                   bottle contents weight 50 g.

              3.7.2.3          0.250 ppb Standard

                     For the 0.250 ppb std, follow steps 3.7.2.2.1 through 3.7.2.2.4. In step
                     3.7.2.2.5, substitute 125 jiL of the 100 ppb working standard  using a new
                     1000 jiL precleaned tip.  Follow steps 3.7.2.2.6 through 3.7.2.2.7.
                                           2-484

-------
                                               SOP for Analysis of Sediment for Total Mercury
                                                    Using the Cold Vapor Technique with the
 Volume 2, Chapter 2                                     Leeman Labs, Inc. Automated Mercury
	System
                                        2-485

-------
SOP for Analysis of Sediment for Total Mercury
Using the Cold Vapor Technique with the
Leeman Labs, Inc. Automated Mercury System	Volume 2, Chapter 2

              3.7.2.4         0.500 ppb Standard

                     For the 0.500 ppb std, follow steps 3.7.2.2.1 through 3.7.2.2.4. In step
                     3.7.2.2.5, substitute 250 jiL of the 100 ppb working standard using a new
                     1000 jiL precleaned tip. Follow steps 3.7.2.2.6 through 3.7.2.2.7.

              3.7.2.5         1.00 ppb Standard

                     For the 1.00 ppb std, follow steps 3.7.2.2.1 through 3.7.2.2.4.  In step
                     3.7.2.2.5, substitute 500 jiL of the 100 ppb working standard using a new
                     1000 jiL precleaned tip. Follow steps 3.7.2.2.6 through 3.7.2.2.7.

              3.7.2.6         2.00 ppb Standard

                     For the 2.00 ppb std, follow steps 3.7.2.2.1 through 3.7.2.2.4.  In step
                     3.7.2.2.5, substitute 1000 jiL of the 100 ppb working standard using a new
                     1000 jiL precleaned tip. Follow steps 3.7.2.2.6 through 3.7.2.2.7.

3.8    Reagents Needed for Preparation of Sediment Samples using Protocol PRP7471

       3.8.1   50% Aqua Regia (3:1 Hydrochloric :  Nitric)

              3.8.1.1         Use the glass container that previously held the Suprapure
                             hydrochloric acid, this is supplied with graduations. Rinse three
                             times with MSQ to eliminate any residual contamination.

              3.8.1.2         To this glass container add 400 mL of MSQ water.

              3.8.1.3         To the glass container add 300 mL of Leeman Labs hydrochloric
                             acid to the bottle.

              3.8.1.4         Finally add to the bottle 100 mL of Leeman Labs nitric acid.

              3.8.1.5         Carefully transfer solution to instrument bottle.

       3.8.2  Potassium Permanganate Solution

              3.8.2.1         Rinse a 100 mL precleaned finely graduated cylinder three times
                             with 5% reagent grade nitric acid followed by three rinses of
                             MSQ.

              3.8.2.2         Dispose of the instrument bottle's contents if applicable and rinse
                             three times with MSQ.

              3.8.2.3         Add 40mLs of the Leeman Labs potassium permanganate
                                           2-486

-------
                                             SOP for Analysis of Sediment for Total Mercury
                                                  Using the Cold Vapor Technique with the
 Volume 2, Chapter 2                                   Leeman Labs, Inc. Automated Mercury
	System

                         solution to the rinsed 100 mL graduated cylinder.

                                  WARNING

  Potassium permanganate is a strong oxidizer. Keep it from contact with clothing or
  combustible materials.  Avoid contact with eyes or skin. Avoid breathing the dust.
  Handle solid chemical under a fume hood. Handle it while wearing personal
  protection gear.

         3.8.2.4         Transfer to the instrument bottle.

         3.8.2.5         Add 760 mL of MSQ to the bottle.

         3.8.2.6         Swirl the contents of the bottle to ensure a mixed solution.

         3.8.2.7         Attach to the instrument.

  3.8.3  Hydroxylamine Sulfate Solution

         3.8.3.1         Rinse a 100 mL precleaned  finely graduated cylinder three times
                         with 5% reagent grade nitric acid followed by three rinses of
                         MSQ.

         3.8.3.2         Dispose of the instrument bottle's contents if applicable and rinse
                         three times with MSQ.

                                      WARNING

  Hydroxylamine sulfate is an eye, skin, inhalation,  and ingestion hazard.  It will
  cause skin irritation and may be absorbed through the  skin.  Always wear eye
  and skin protection.

         3.8.3.3         Add 96mL of the Leeman Labs Hydroxylamine Sulfate solution
                         to the lOOmL graduated cylinder.

         3.8.3.4         Transfer the Hydroxylamine Sulfate to the instrument bottle and
                         continue to add 704 mL of MSQ to the instrument bottle.

         3.8.3.5         Swirl bottle to ensure a mixed solution.

         3.8.3.6         Attach bottle to instrument.
                                      2-487

-------
SOP for Analysis of Sediment for Total Mercury
Using the Cold Vapor Technique with the
Leeman Labs, Inc. Automated Mercury System	Volume 2, Chapter 2

                                         NOTE

       This is a modification of the protocol that calls for full strength concentrations
       of Potassium Permanganate and Hydroxylamine Sulfate. Full strength
       concentrations of these chemicals were attempted and the fumes generated
       were very strong.  There was no difference in instrument performance in using
       some higher concentrations of reagents in the protocol. Both gave acceptable
       recoveries on the SRM material and similar instrumental intensities.

4.0   Microwave Sample Preparation

4.1     Preparation of Teflon Digester Vessels

       Teflon digester vessels are cleaned following the procedure in section 4 of LLRS-MET-
       SOP-010. An alternate method proposed by the manufacturer to increase the life of the
       digestion vessels is under consideration.  It may not be used until it is verified that blanks
       are equally low for the two methods. Upon verification, the following alternate steps may
       be substituted for steps 4.4 through 4.7 in LLRS-MET-SOP-010. Steps are taken directly
       from CEM manual with slight variations to fit constraints of laboratories (Oilman 1988).

       4.1.1   Add 20 mL of concentrated nitric acid to the digestion flask. Place the safety disk
              on the vessel and tighten finger tight only. Place the vessel in the turntable, and
              attach a venting tube.

       4.1.2   Repeat step until the turntable contains 12 vessels.

       4.1.3   Turn the MDS-8ID exhaust onto the maximum fan speed.  Ensure the turntable is
              rotating.

       4.1.4   Program the instrument for five minutes and 100% power.  Depress the start key
              and  allow the acid to heat.

       4.1.5   Allow the acid to cool to room temperature and manually vent each vessel. Open
              vessels and pour the acid into an appropriate waste container.

       4.1.6   Rinse the vessels three times with MSQ water and allow them to dry in a clean
              area.
                                          2-488

-------
     Volume 2, Chapter 2
 SOP for Analysis of Sediment for Total Mercury
      Using the Cold Vapor Technique with the
        Leeman Labs, Inc. Automated Mercury
	System
4.2    Extraction of Samples

       The maximum weight of samples extracted in CEM vessels without venting and still
       obtaining acceptable recoveries for all metals has been two grams. For mercury, an
       average weight of O.SOOg has been used. The total volume is 25mL in 10% Seastar nitric
       acid. Follow steps 5.1.1 through 5.1.11 in LLRS-MET-SOP-010 for preparation of
       sediments

5.0    AUTOMATED DIGESTER (AP200)  SAMPLE PREPARATION

5.1    Daily Instrument Setup

       5.1.1  Clean autosampler rails with isopropyl  alcohol.

       5.1.2  Lubricate rail s with oil daily

                                        NOTE

       Only if reagents are low or instrument was in shutdown will following steps
       need to be followed (5.1.3 through 5.)

       5.1.3  Open cover and carefully  disconnect each bottle separately from the interior of the
             instrument.

       5.1.4  Rinse each bottle three times with MSQ

       5.1.5  Fill bottles numbered 1, 5, 6 with MSQ up to the 800ml mark.

       5.1.6  Prepare and fill bottle #3 with the Potassium Permanganate Solution.

       5.1.7  Prepare and fill bottle #4 with Hydroxylamine Sulfate Solution

       5.1.8  Carefully prepare and fill bottle #2 with the 3:1 Aquaregia solution

       5.1.9  Check conditions of all fittings, caps and bottles

       5.1.10 Close cover and pressurize system (turn gas on).

       5.1.11 Set gas pressure = 20 psi

                                      CAUTION

                                  Do not exceed 25 psi
                                         2-489

-------
SOP for Analysis of Sediment for Total Mercury
Using the Cold Vapor Technique with the
Leeman Labs, Inc. Automated Mercury System	Volume 2, Chapter 2

       5.1.12  Turn instrument on

       5.1.13  Turn computer and monitor on.

       5.1.14  Turn printer on and press the online button.

       5.1.15  After the computer has booted up, at the C: prompt, type apps.

       5.1.16  Follow instructions on p4-4 of manual revision c.

5.2    Software and Instrument Setup

       Follow the instruction in system startup section 4-3, What follows is a summary.  The user
       should review the following sections prior to analyses.


                     References to AP200 Manual

                     Section 3- System Testing
                     Section 4- System Operation
                     Section 5- Routine Maintenance
                     Section 6- Troubleshooting

       5.2.1   Select Protocal and Get PRP7471, defines method, (sec 4-4 of manual)

       5.2.2   Check reagent pressure of system, ensure it is > then 5.5psi and within 6.5.
              (section 2-10 of manual)

       5.2.3   Go to menu, Fl, select Utility, select Diagnostics

       5.2.4   Select Reagent Pressure

       5.2.5   Run the change rinse solution macro, @CF£RTNS

       5.2.6   Rinse the autosampler rinse tray 3-4 times with MSQ

       5.2.7   Replace the rinse tray

       5.2.8   Run the wake up macro @WAKEUP (section 4-3 of manual)

       5.2.9   Check centering of autosampler tip over cups, adjust if necessary (section 3-1)

       5.2.10  Check precision of dispenser IX/week (section 3-3), or if there is a pressure
              change of .2 psi in system at step 14.3.3
                                          2-490

-------
                                                  SOP for Analysis of Sediment for Total Mercury
                                                       Using the Cold Vapor Technique with the
     Volume 2, Chapter 2                                   Leeman Labs, Inc. Automated Mercury
	System

       5.2.11  Set up autosampler sequence, and start finish sequence  (section 4.6)

5.3    Preparation for Sample Digestion

       5.3.1   Soak the 45mL standard cups and sample cups in  1:1 HNO3 for at least 2.0 hours.

       5.3.2   Recycle the acid from the cups back into the Suprapure Acid bottle.

       5.3.3   Rinse each cup three to six times with MSQ and allow to air dry.

5.4    Preparation of Standards for Automated Digestion

       5.4.1   Prepare an EDP pipettor with a precleaned lOOOul tip.

       5.4.2   Using the electronic scale, tare the beaker, and then the standard cup within the
              beaker.

       5.4.3   Record the weight of the empty cup on the extraction sheet.

       5.4.4   Under the  laboratory  hood, pipette 5mls of the 0 ppb std to the cup.

       5.4.6   Weigh the cup to confirm volume delivered is 5mls, and adjust accordingly with
              the EDP pipettor.

       5.4.7   Repeat steps 16.2 - 16.5 to extract multiple 0 standards.

       5.4.8   Repeat steps 16.1 to 16.5 for the 0.125ppb std, 0.250ppb std, O.SOOppbstd.
              l.OOppb std, and 2.00ppb std.  Each time use a new precleaned lOOOul tip.

       5.4.9   Rinse polyethylene vapor covers 3x with MSQ and place over sample cups.

       5.4.10  Snap in place aluminum guard over polyethylene vapor  barrier.

5.5    Preparation of Sediment Samples for Automated Digestion

                                          Note

       Please read the instruction manual for the Mettler Analytical Balance before
       proceeding.

       5.5.1   Rinse a precleaned teflon spatula with 5% nitric acid.

       5.5.2   Rinse the spatula three times with MSQ.
                                          2-491

-------
SOP for Analysis of Sediment for Total Mercury
Using the Cold Vapor Technique with the
Leeman Labs, Inc. Automated Mercury System	Volume 2, Chapter 2

       5.5.3   Use the analytical balance in the Biology Lab, Mettler 21 GOT

       5.5.4   Tare the lOOmL polyethylene beaker provided.

       5.5.5   Add the 45mL sample cup to the beaker and tare the scale again.

       5.5.6   Record the weight of the sample  cup on the digestion sheet.

       5.5.7   Carefully open the Whirlpak bag  of sediment.

       5.5.8   Open the Whirlpak bag partially,  so only a small circular opening exists (about the
              size of a nickel).  Caution: if the bag is open all the way, dust from the bag tends to
              migrate upwards.

       5.5.9   Using the precleaned teflon spatula, carefully scoop an aliquot of sample = 0. lOOg.

       5.5.10  Transfer this aliquot to the sample cup.  Note: sand like samples will require  a
              very small aliquot, and silty samples will require a larger aliquot.

       5.5.11  Record the weight of the sample in the most significant digits available  on the
              extraction log sheet.

       5.5.12  Place the cup in its designated slot in a sample rack.

       5.5.13  Zero the scale with the polyethylene beaker on it.

       5.5.14  Rinse the teflon spatula with MSQ water three times.

       5.5.15  Wipe teflon spatula dry with a fresh clean wipe square between samples.

       5.5.16  Repeat steps 16.6.5 till  16.6.15 for the desired amount of samples, usually
              fourteen.

       5.5.17  Load dummy cups into any space not occupied by  actual sample.

       5.5.18  Rinse polyethylene vapor covers  3x with MSQ and place over sample cups.

       5.5.19  Snap in place aluminum guard over polyethylene vapor barrier.

5.6    Initiation of the Digestion Procedure

       5.6.1   Load autosampler racks into PS200

       5.6.2   Confirm autosampler start to finish sequence is correct.
                                           2-492

-------
                                                 SOP for Analysis of Sediment for Total Mercury
                                                     Using the Cold Vapor Technique with the
     Volume 2, Chapter 2                                   Leeman Labs, Inc. Automated Mercury
	System

       5.6.3  Go to Main Menu, Fl, press MACRO key, and type in @PRP7471

       5.6.4  Method will begin to run at this point.

       5.6.5  User will be prompted the following: Wait before reducing with KMNO4,
             answer Y

                                      CAUTION

User will be prompted near end of 5 hr procedure: Add Hydroxylamine Sulfate, answer Y.
Method will not proceed without this input

5.7    Shutdown of AP200

       5.7.1  After protocol is completed, wipe down bath to remove any remaining water.

       5.7.2  Run an @APNAP two or three times to clean the dispenser with water.

       5.7.3  Remove chemicals and replace reagents with MSQ if it will not be run for an
             extended period, and run an @CLEAN twice.  This cleans not only the dispenser,
             but all reagent lines.

6.0    Automated Analysis of Digested Extracts

6.1    Preparation of the Instrument

       6.1.1  Preparation of Drying Tube

             6.1.1.1 Rinse a quartz drying  tube three times with MSQ, followed by a dilute
                     rinse of Liquinox, if it previously contained perchlorate.

             6.1.1.2 Rinse several times with MSQ to  eliminate Liquinox residuals.

             6.1.1.3 Allow tube to air dry.

             6.1.1.4 Rinse the teflon spatula three times with MSQ and dry with a fresh clean
                     wipe.

             6.1.1.5 Gently place a small plug of quartz wool into one end of drying tube.

             6.1.1.6 Carefully pour the Leeman Labs Magnesium Perchlorate into     the
                     plugged drying tube.  Try to fill with coarse grained perchlorate.  Do not
                     overfill the drying tube.  Overfilling will cause the drying tube to become
                     blocked more easily once it becomes moistened by the gas stream. When
                                         2-493

-------
SOP for Analysis of Sediment for Total Mercury
Using the Cold Vapor Technique with the
Leeman Labs, Inc. Automated Mercury System	Volume 2, Chapter 2

                      filled, the perchlorate should be able to move within the tube when gently
                      moving the drying tube from side to side.
                                           2-494

-------
                                             SOP for Analysis of Sediment for Total Mercury
                                                  Using the Cold Vapor Technique with the
 Volume 2, Chapter 2                                    Leeman Labs, Inc. Automated Mercury
	System

                                   WARNING

  Magnesium perchlomte is moderately toxic and a strong oxidizing material.  It
  is a dangerous fire and explosion risk in contact with organic materials. It is
  an inhalation hazard, and contact with skin or eyes can cause irritation.  Work
  with it in a fume hood while wearing skin and eye protection.

  6.1.2  Autosampler Tray Rinse with 10% Hydrochloric Acid

         6.1.2.1  Rinse the autosampler rinse tray three times with MSQ.

         6.1.2.2  Fill the tray with the 10% hydrochloric rinse prepared in step 3.1.

  6.1.3  Tin Chloride Rinse

         6.1.3.1  Rinse the tin chloride bottle out three times with MSQ.

         6.1.3.2   Fill it with the 10% tin chloride prepared in  step 3.3.

  6.1.4  Check Tubing Condition and Adjust to Appropriate Tension

         6.1.4.1  Check condition of tubing for flattening,  abrasion or other signs of wear.
                 If flattened, replace it.

         6.1.4.2  Adjust tension on clamps to a halfway point. The sample line should be
                 halfway minus one notch.

                                   CAUTION

  Do not over tighten clamps.  Too much tension  will cause tube flattening and a
  decrease in overall sensitivity.

  6.1.5  Clean and Oil Autosampler Rails

         6.1.5.1  Using a clean wipe or clean a paper towel, wipe the autosampler rails with
                 isopropyl alcohol.

         6.1.5.2  Place a small amount of oil on bottom of each rail.

         6.1.5.3  Complete a high stress maintenance cleaning monthly.
                                      2-495

-------
SOP for Analysis of Sediment for Total Mercury
Using the Cold Vapor Technique with the
Leeman Labs, Inc. Automated Mercury System	Volume 2, Chapter 2

       6.1.6  Warmup Period

             6.1.6.1  Allow the instrument to warmup for at least one hour time before
                     analyses.  If the instrument was in SHUTDOWN mode for an extended
                     period allow it to warm up for several hours.

             6.1.6.2  Software  commands for warmstrt or coldstrt are found on page 16 of the
                     manual.

             6.1.6.3  Perform a COLDSTRT or WARMSTRT

       6.1.7  Optimize Optics with an Aperture Test

             6.1.7.1  Go to Main Menu, Fl

             6.1.7.2  Select Diagnostics, Select Aperture Test

             6.1.7.3  Unscrew screw which is furthest out until the minimum absorbance is
                     obtained.  An acceptable value is 0-100.

             6.1.7.4  Select Test Optics from Diagnostics menu, and confirm gain, or intensities
                     are in the range of 500000-1100000 (= to voltage on lamp).

             6.1.7.5  Difference between the two beams must be less than 100,000.

6.2    Software Setup for Routine Analysis

       Consult Leeman Labs Automated Mercury Analyzer Manual pp 17-29 for guidance on the
       software setup. What follows are the  software and instrumental parameters used to date.
       Complete the following steps before analyses.

       6.2.1  Establishing a Protocol = Method file that Contains All Instrumental Parameters

             6.2.1.1  From the  Main Menu, select Protocol  and then select Get.

             6.2.1.2  To create a new Protocol, enter its name.

             6.2.1.3  Suggested protocol naming is as follows: YYMMDD (Year, Year,
                     Month, Month, Day, Day. Example 960916F = September 16, 1996.

                                         Note
                               Limitation is 8 Characters

             6.2.1.4  Computer will guide you through prompts (PS200 Manual pp. 18-19).
                                         2-496

-------
                                              SOP for Analysis of Sediment for Total Mercury
                                                   Using the Cold Vapor Technique with the
 Volume 2, Chapter 2                                    Leeman Labs, Inc. Automated Mercury
	System

   6.2.2  Creating a Folder=Data File

         6.2.2.1 Press Fl, to be at the Main Menu.

         6.2.2.2 Select Data output and select Open Folder.

         6.2.2.3 Type in a folder name, suggested name = same name as protocol to avoid
                 confusion.

         6.2.2.4 Follow instructions on page 19 of the manual.

   6.2.3  Entering Instrumental Parameters

         6.2.3.1 Follow instructions in PS200 instrument manual on pp 20-21.

         6.2.3.2 The following are parameters used to date:

                 Integrations:       1
                 Uptake Time:      10
                 Weight:           Y
                 Dilution:          Y
                 Percent Recovery:  N
                 On/Off:    10  (higher integration time generates a nonlinear curve)
                 Flow Rate:        0.30L/min

   6.2.4  Standard Concentration Calculations

         For those standards that are prepared with the automated digestion system, the
         appropriate concentration to be used for keying in standard concentration data
         must be calculated. Do not do this for microwave digested samples.

         6.2.4.1 Tare a 100ml polyethylene beaker on the electronic scale.

         6.2.4.2 Weigh a sample cup.

         6.2.4.3 Record the weight on the extraction log, and zero the scale.

         6.2 A A Repeat 6.1.1 through 6.1.3 for all sample cups.
                                      2-497

-------
SOP for Analysis of Sediment for Total Mercury
Using the Cold Vapor Technique with the
Leeman Labs, Inc. Automated Mercury System	Volume 2, Chapter 2

              6.2.4.5 Apply the following formula to calculate the final concentration of the
                     extracted standards:

                                     c = r * v /v
                                     ^2  M   v 1'  V2 •>

                                         where,

                         Cj is the prepared  standard concentration,
                         C2 is the final concentration in the standard,
                         Vj is the volume  of prepared  standard used,
                          and V2 is the total volume of the extract

                     For example if 5 mL of 0.500 ppb standard was added to the digestion
                     tube and the final volume of the extract is 43 mL, the resulting standard
                     concentration is as follows:

                             C2 = 0.500 ppb * 5 mL/43 mL = 0.058 ppb.

              6.2.4.6 Record this concentration on the extraction log.

       6.2.5   Entering Standard Concentrations

              6.2.5.1 Follow instructions on  p. 21 of the manual. Enter units in ppb, not ppm
                     (i.e. 0.500 not 0.00050).

              6.2.5.2 Calculations are only carried out  to three decimal points, 0.00050 will be
                     truncated to 0.000.

              6.2.5.3 Do not enter terms of units, i.e. ppb,  ppm. The final calculation will be in
                     ug/g.

              6.2.5.4 The following are ranges of standards used for analysis.  These ranges
                     have been successful in bracketing low level samples.

       Microwave Digestion Standards (ppb)       Automated Prepared Standards (ppb)
                                                (Dependent upon total volume of extract.
                                                See section 6.2.4)

                     0.000                             0.0000
                     0.250                            -0.0140
                     0.500                            -0.0280
                     1.000                            -0.0570
                     2.000                            -0.1140
                     4.000                            -0.2280
                                          2-498

-------
                                                 SOP for Analysis of Sediment for Total Mercury
                                                      Using the Cold Vapor Technique with the
     Volume 2, Chapter 2                                   Leeman Labs, Inc. Automated Mercury
	System

       6.2.6  Reset Calibration Intensity Data

             Follow instructions on page 22 of manual.

       6.2.7  Autosampler Rinse Time

             6.2.7.1 Follow instructions on page 23 of manual.

             6.2.7.2 Use a rinse time of 60, not 50 seconds. This is the rinse time between
                     samples in the analyses mode.

       6.2.8  Autosampler Rack Entry

             6.2.8.1 For basic entry information, refer to page 25 of the manual.  For extended
                     information  on macros and advanced command see reference section A-
                     B-7.

             6.2.8.2 What follows is an example of an autosampler rack file. It was used for
                     analyses of Green Bay sediment. Prepare the file before analyses.

                                         NOTE

       Actual sample weight must be multiplied by 1000 to obtain results in ug/g.
       Total volume = extraction volume * dilution factor.

       1 OPPB                1.000
       2GII25F102SQ*       323.00            250.00              SRM  diluted IQx
                                                                   extracted in 25mls
       3 GB88-71             280.00            125.00              Sample diluted 5x
                                                                   extracted in 25mls
       4GB89-73             241.00            125.00
       5 GB89-74             5.700              125.00

                     A microwave digested SRM will need to be diluted 10 to 20 times
                     depending upon the weight of the sample to be within range of standards.
                     Certified value for SRM2704 = 1.47 ug/g.  For a sample that has a
                     extraction weight of 0.250g in 25mL, (1.47*.250)/25=0.018375
                     ppm=18.37 ppb.  This is diluted 20X = .918 ppb

6.3    Analysis of Extracts

       6.3.1  Filling Autosampler Cups

             6.3.1.1 Recycle the 50% acid rinse used in the autosampler cups.  Place it in the
                                         2-499

-------
SOP for Analysis of Sediment for Total Mercury
Using the Cold Vapor Technique with the
Leeman Labs, Inc. Automated Mercury System	Volume 2, Chapter 2

                     glass SUPRAPURE acid bottle.

              6.3.1.2 Rinse each cup 5-10 times with MSQ.

              6.3.1.3 Allow cups to air dry

              6.3.1.4 Fill each standard cup with its designated standard.

              6.3.1.5 Using two separate EDP pipettors, prepare a clean lOmL tip and lOOOuL
                     tip. Rinse each tip three times with 5% nitric acid followed by three rinses
                     ofMSQ.

              6.3.1.6 For samples and SRMS requiring dilution, dilute to at least half their
                     capacity of the autosampler tips (6mL).  First add the required volume of
                     diluent with the lOmL tip and then the required amount of the sample.  An
                     SRM will need to be diluted 10 to 20 times depending on weight of
                     sample. For dilutions of auto-digested samples use 0 ppb standard that
                     has undergone digestion. For microwave digested samples use 10%
                     Seastar nitric acid.

              6.3.1.7 Mix the sample 3-5 times with  the lOOOuL tip.

              6.3.1.8 Use a new precleaned lOOOuL tip for each sample.

       6.3.2   Calibrate the Instrument

              Calibrate the instrument using the Macro CAL245 (p. 26 in PS200 instrument
              manual). Use a 5-point calibration curve that includes a zero standard.  If an
              acceptable correlation coefficient is obtained (0.995) and a standard's intensity is
              within a the range expected, continue with SRM analyses. See Appendix A for
              historic performance of the instrument

       6.3.3   Analyze the SRM

              Analyze the SRM. Refer to page 25 in  manual for autosampler start to finish
              sequence and reference section A-B 7.

       6.3.4   Check Standards

              6.3.4.1 Run check standards every 10 samples to ensure the instrument has not
                     drifted from its calibration range.
                                          2-500

-------
                                                 SOP for Analysis of Sediment for Total Mercury
                                                      Using the Cold Vapor Technique with the
     Volume 2, Chapter 2                                   Leeman Labs, Inc. Automated Mercury
	System

             6.3.4.2 Acceptable Check Standard Ranges are:
                                  Microwave Digestion        Automated Digestion

                                  0.250ppb = 15%             ~0.014ppb = 15-20%
                                  0.500ppb=10%              -0.028 ppb  = 15%
                                  1.00ppb=10%               ~0.058ppb = 10%
                                  2.00ppb=10%               -0.115 ppb  = 10%

             6.3.4.3 Refer to reference section c-11 for more information.

                                       CAUTION

       If check standards fail, recalibrate the instrument. Do not use update slope or
       intercept.

6.4    Data File Preparation

       Refer to the reference sections D-l, E-l-5, R-3, and R-5 for preparation of post-run data
       and computer files in the AP200 Manual.  These data references apply to digested and
       samples prepared on the Automated Digester.

6.5    Instrument Shutdown

       6.5.1  Dispose of 10% Hydrochloric acid rinse in an appropriate container.

       6.5.2  Rinse the tray out three times with MSQ.

       6.5.3  Transfer remaining tin chloride to teflon bottle in which it was earlier prepared.

       6.5.4  Rinse out tin chloride bottle three times with MSQ.

       6.5.5  Fill autosampler tray and tin chloride bottle with MSQ, flush for ten minutes.

       6.5.6  Use OVERNITE, or SHUTDOWN modes to shutdown the instrument.

       6.5.7  If using OVERNITE MODE, check condition of drying tube, to ensure it is not
             saturated with moisture.

       6.5.8  Repack a new drying tube if necessary.
                                          2-501

-------
SOP for Analysis of Sediment for Total Mercury
Using the Cold Vapor Technique with the
Leeman Labs, Inc. Automated Mercury System	Volume 2, Chapter 2

7.0   Suggestions for Successful Analyses

7.1     Allow the autosampler cups and standard cups to rest in 50% nitric acid for at least two hours
       before analyses.

7.2     For best results prepare tin chloride and standards as described.

7.3     Prepare a loosely packed drying tube daily. After use, dispose of perchlorate in an appropriate
       container. Rinse a drying tube with MSQ and flush lightly with Liquinox. Rinse several times to
       eliminate any residual Liquinox.

7.4     Periodically check tin chloride line and liquid gas separator for any blockage.

7.5     Check at seals of teflon tubing of drying tube connection for gas leaks.

7.6     Change tin chloride line and sample line weekly or after four days of continuous use.

7.7     Change drain tubing every two weeks as needed.

7.8     Clean autosampler rails with isopropyl alcohol weekly and oil rails daily. If not sufficiently
       lubricated, the autosampler arm  will encounter snags or stops.

7.9     Biweekly calibrate the autosampler tip to ensure it is picking up more  than three milliliters.

8.0   Literature Cited

Oilman, L. B., 1988.  General Guidelines for Microwave Sample Preparation. Revision of July
   1988.  CEM Corporation, Matthews, NC.

Leeman Labs,  Inc., 1991.  PS200 Automated Mercury Analyzer Set-up  and Operating Manual.
  Version of November 1991.

Leeman Labs,  Inc., 1993.  AP200 AutomatedMercury Preparation System Manual.  Revision C
  (11/20/93).

Rossmann, R. And K.  A. Rygwelski, 1996.  Standard Operating Procedures for the Release of
  Data.  Revision 1 (1/18/96) LLRS-QA-SOP-001.

Rossmann. R., 1992. Standard Operating Procedures for the Routine Review of Data Quality
  and Quantity. Revision 0 (12/2/92) LLRS-QA-SOP-002.

Rossmann, R., 1992. Minimum Analytical Quality Assurance Objectives for U. S. EPA Large
  Lakes Research Station.  Revision 1 (7/28/92) LLRS-QA-001.

Rossmann, R., 1993. Standard Operating Procedures for the Preparation of Materials used for
   Ultra-low Trace Element Analyses. Revision  1 (8/9/93) LLRS-MET-SOP-001
                                          2-502

-------
                                               SOP for Analysis of Sediment for Total Mercury
                                                    Using the Cold Vapor Technique with the
     Volume 2, Chapter 2                                   Leeman Labs, Inc. Automated Mercury
	System

Rossmann, R., 1994. Standard Operating Procedures for the Maintenance of the LLRS Trace
  Metal Laboratories.  Revision 1 (5/4/94) LLRS-MET-SOP-003.


Rossmann, R. and T. Uscinowicz, 1994. Standard Operating Procedures for Analysis of Total
  Mercury in Tissue and Sediment using the Cold Vapor Technique with the Perkin-Elmer
  Model MHS-20 Gold Amalgam System. Revision 1 (5/6/94) LLRS-MET-SOP-010.
                                        2-503

-------

-------
    Volume 2, Chapter 2
 SOP for Analysis of Sediment for Total Mercury
     Using the Cold Vapor Technique with the
       Leeman Labs, Inc. Automated Mercury
	System
APPENDIX: HISTORIC PERFORMANCE FOR STANDARDS ON PS200.

Daily Standard Intensities and Percent Drift for Standards. Italicized dates are check standards.
Date of
Analysis
01/17/96
01/18/96
01/24/96
01/25/96
01/29/96
01/31/96
02/01/96
02/05/96
02/05/96
02/06/96
2/6/96
2/6/96
2/6/96
02/08/96
2/8/96
2/8/96
02/13/9
2/13/96
2/13/96b
2/13/96b
02/15/96
2/15/96
2/15/96b
Begin New Lamp
03/04/96
03/04/96
03/04/96
03/05/96
03/05/96
03/06/96
0.250

11567
15405
14521
12899
12153
13360






15429


13614

13675

13628

12642

32237
31507
29918
22484

22151
Standard Concentration (ppb)
0.500 1.000 2.000
30807
24577
27507
25177
30207
25133
26287
28271
28294
21563
25966
24807
23963
31433
32989
31222
27169
30254
26879
28977
24557

21928

44465
46352
41993
65754
64815
45823
62450
52846
62220
55494
54795
47097
52552
55116
57208
43415
47569
49621
48835
62122
62558
60391
59876
63307
58308
60081
57516
54759
53348

127713
130143
119879
124625
114069
90596
128462
112763
127897
117531
108733
110427
113649
107236
116260
95356
103997
102217
99624
124122
121403
120489
121738
125174
121548
122166
118139
111736
112076

233412
226759
217858
264984
259107
180965
5.000
319499
289297
309450
279308
283705
280385
286800
262538
287149
258884
263870
259967
251241
312746
316603
312282
299635
304689
297642
278297
286801
268327
270167

620712
626601
598765
628674
609113
422997
                                     2-505

-------
SOP for Analysis of Sediment for Total Mercury
Using the Cold Vapor Technique with the
Leeman Labs, Inc. Automated Mercury System
Volume 2, Chapter 2
                             APPENDIX:  Continued.

Daily Standard Intensities and Percent Drift for Standards. Italicized dates are check standards.
Date of
Analysis
03/07/96
03/07/96
03/12/96
03/14/96
03/14/96
03/14/96
03/14/96
03/15/96
03/15/96
03/19/96
03/22/96
03/26/96b
03/28/961
03/28/962
03/29/96
03/29/96
03/29/96
03/29/96
04/03/96
04/09/96
04/09/96
04/09/96
04/11/96
1 Old Tubing
2 New Tubing
Standard Concentration (ppb)
0.250
21132
19282

23058
24152
24206
24008
25333
23274

23430
22930
23663
27619
22912

21975

20843
20494
42012
40964
21043


0.500
44232
39037
37940
46864
47468
47189
47491
48936
48272
35727
43189
45156
44270
53606
47100

44362
44504
42173
41875


42323


1.000
95743
79923
79106
96048
95645
96993
94850
98592
98146
81247
90404
84083
79330
109573
96970

90593

85748
83552

81646
89779


2.000
194749
202894
166230
192849
188087
188049
187538
195526
197376
178190
194519
188764
184093
212636
194573
191256
196257

167910
169354
167949
160894
178299


5.000
478754

411516
457716
469761
454648
453447
485709
473520
469406
493374
465947
457756
523307
483750
474460


429912
419310

401227
444557


                                         2-506

-------
     Volume 2, Chapter 2
 SOP for Analysis of Sediment for Total Mercury
      Using the Cold Vapor Technique with the
        Leeman Labs, Inc. Automated Mercury
	System
                             APPENDIX:  Continued.
Daily Standard Intensities and Percent Drift for Standards Prepared with Automated Digestion
System.  Italicized dates are check standards.
Date of
Analysis
06/1 1/96
06/13/96
06/18/96
07/03/96
07/09/96
07/12/96a
07/12/96b
07/16/96
07/17/96
07/24/96
07/25/96
08/01/96
08/15/96
08/16/96
09/18/96
-0.014








927
1175
980
600
895
1069
816
Standard Concentration (ppb)
-0.028 -0.056 -0.114





1937
2162
1867
1973
2124
1923
1728
1678

1819
4278
4275
4138
5043

4098

4308
4457
4146
4492
3989
3910
4189
360
8331
8689
8979
8057
8369
8088
8118
8488
8454
8043
7546
7908
7977
8468
7382
-0.228

17993
17995
15743
18117
17318
17009
17285
16962
16190

17072
15964

15077
                                         2-507

-------