Guidance on the Documentation and Evaluation of
          Trace Metals Data Collected for
     Clean Water Act Compliance Monitoring
                        Draft
                      July 1996
          U.S. Environmental Protection Agency
                    Office of Water
            Office of Science and Technology
         Engineering and Analysis Division (4303)
                     401 M St. SW
                 Washington, DC 20460

-------
Acknowledgments
This  guidance  was  prepared  under  the  direction  of  William  A.  Telliard  of  the  Engineering
and   Analysis   Division  (BAD)  within  the  U.S.   Environmental  Protection  Agency's  (EPA)
Office  of  Water  (OW),   Engineering  and  Analysis   Division  (BAD).     This   guidance  was
prepared  under   EPA  Contract   68-C3-0037   by  the  DynCorp  Environmental   Programs
Division with assistance from Interface, Inc.
Disclaimer

This  document  has  been  reviewed  and  approved  for  publication  by  the  Analytical  Methods
Staff  within  the  Engineering and  Analysis  Division  of the  EPA  Office  of Water.   Mention
of  trade  names  or  commercial  products  does  not  constitute  endorsement  or  recommendation
for use.
Further Information

For further information, contact:

       William A. Telliard, Chief
       Analytical Methods Staff
       Engineering and Analysis Division
       U.S. Environmental Protection Agency
       401 M Street
       Washington, DC 20460
       Phone:  202-260-7134
       Fax:    202-260-7185


Requests for additional copies should be directed to:

       US EPA NCEPI
       11029 Kenwood Road
       Cincinnati, OH 45242
       513-489-8190
                                                                               Draft, July 1996

-------
                                                                              	Chapter 1

                                                                                       Introduction

        Numerous   organizations,  such  as  state   pollution  control  agencies,  health  departments,
local  government   agencies,  industrial   dischargers,  research  facilities,  and   federal  agencies   (e.g.,
EPA,  USGS),  collect  data  on effluent and   ambient  metal  concentrations  for  use  in  a  variety
of   applications,   including:     determining   attainment   status   for  water  quality  standards,
discerning   trends   in  water  quality,   estimating   effluent   concentrations   and  variability,
estimating  background  loads  for   total  maximum  daily  loads   (TMDLs),   assessing   permit
compliance,  and  conducting research1.    The  quality  of  data  used  is  an  important issue,  and,
in particular, the  quality  of trace level metals  data  may  be  compromised  due  to  contamination
during  sampling, filtration,  storage,  and  analysis.    In  fact, one  of the  greatest  obstacles  faced
by  laboratories  attempting  trace   metals  determinations   is  the   potential   for   contamination  of
samples   during the   sampling  and  analytical   processes.    Trace   metals   are  ubiquitous  in  the
environment,  and   samples  can  readily  become  contaminated   by  numerous  sources,  including:
metallic  or   metal-containing   labware,  metal-containing  reagents,   or  metallic   sampling
equipment;   improperly  cleaned  and   stored   equipment;  and  atmospheric   inputs  such  as  dirt,
dust, or other participates from exhaust or corroded structures.

        The  measurement  of  trace  metals  at  EPA  water  quality  criteria  (WQC)  levels  has  been
spurred  by   increased  emphasis  on   a water  quality-based  approach  to   the   control  of   toxic
pollutants.    Current   ambient  WQC   levels2   for  trace  metals   require  measurement  capabilities
at levels  as  much  as 280  times lower than those  levels  required  to  support  technology-based
controls   or   achievable  by  routine   analyses  in environmental laboratories.    Also,  recent  USGS
and  EPA  studies  strongly  indicate   that   rigorous  steps   must  be   taken   in  order  to  preclude
contamination during the collection and analysis of samples for trace metals.

        In  order to   ensure  that  the   data  collected  for   trace  metals  determinations  at ambient
water  quality  criteria  levels   are  valid  and   not  a  result   of  contamination,   rigorous   quality
control   (QC)   must   be   applied   to  all  sample   collection,  preparation,   and  analysis  activities.
EPA   has  published   analytical  methods   (1983,  1991)   for   monitoring  metals   in  waters   and
wastewaters,   but    these  methods   are   inadequate   for  the  determination   of   ambient
concentrations   of  metals  in ambient  waters   due   to  the  lack  of some  or  all  of  the  essential
quality  control  and  handling  criteria.    This  prompted  the  Engineering  and   Analysis  Division
(BAD)  to   develop  new  sampling   and  analytical   methods  that  include   the   rigorous   sample
handling  and  quality  control   procedures   necessary  to   deliver   verifiable  data  at  WQC  levels.
The  new  sampling  method is  entitled,  Method  1669:  Sampling Ambient  Water for Determination  of Trace
Metals  at  EPA  Water Quality Criteria  Levels  ("Method  1669").  The  new  analytical  methods  include
Methods  1631,  1632,  1636,  1637,  1638,  1639, and  1640  ("the  1600  Series  Analysis Methods").
Many  of  these  analysis   methods  were   developed   by   supplementing   existing  EPA   methods
        Prothro, M., Acting Assistant Administrator for Water, Memorandum to Water Management Division Directors and
        Knvironmental Services Division Directors, Oct. 1, 1993.


        "Water Quality Standards; Establishment of Numeric Criteria for Priority Toxic Pollutants; States' Compliance" (also referred to
        as "The National Toxics Rule"), 40 CFR Part 131, (57 PR 60848, December 22,1992), and Stay of Federal Water Quality Criteria
        for Metals, 40 CFR Part 131 (60 PR 22228).
Draft, July 1996

-------
with  additional   quality  control   and  sample  handling  requirements;  others  are  new  methods
that are based on newly developed analytical procedures.

        Appropriate  quality  assurance  (QA)  and  quality  control  (QC)  procedures  are  the  key
to  producing   precise  and   accurate   data  unbiased  by  contamination.    Examination  of   trace
metals  data without data  from   blanks  and   other   QC  analyses  yields  little  or  no  information
on  whether sample  data are  reliable.   Data  quality  must  be  documented  through the  use  of
blanks   (both   field  and  laboratory  blanks),   standards,  matrix   spike/matrix  spike  duplicates,
and field  duplicates,  as well as  other QC  analyses.   The results  of all  QC  procedures must  be
included  in  the  data  reporting  package along with the  sample  results  if data  quality  is  to  be
known.

        The  remainder  of  this   document  contains  guidance  that   is  intended   to  aid  in  the
review   of  trace  metals   data   submitted  for  compliance   monitoring   purposes   under   the
National   Pollutant  Discharge   Elimination   System   (NPDES)   when  these    data  are   collected
in  accordance  with  Method  1669 and analyzed  by  the 1600 Series  Analysis  Methods.    Chapter
2   of  this  document  outlines  the  data  elements  that  must  be reported  by   laboratories   and
permittees  so  that  EPA  reviewers  can  validate   the  data.     Chapter   3  provides   guidance
concerning the review  of  data  collected  and  reported  in accordance  with   Chapter  2.    Chapter
4 provides a  Data Inspection  Checklist that can be  used  to  standardize  procedures   for  documenting
the findings of each data inspection.

        The  guidance   provided  in  these   chapters   is  similar in  principle  to  the  data reporting
and review guidance provided in EPA's Guidance on Evaluation,  Resolution,  and Documentation of Analytical
Problems Associated  with  Compliance   Monitoring  (EPA  821-B-93-001),  but  has  been  specifically  adapted
to reflect particular concerns related to the evaluation of data for trace metals.

        This guidance  is  applicable  to  the  examination  of  recently  gathered  trace  metals  data
and  to  historical  data in  existing  EPA  databases.    It  should  be  noted,   however  that  some
qualification  of  historical data  may  be required  before  these  data  can  be  included  in  current
databases.  A draft User's Guide to  the Quality Assurance/Quality Control Evaluation Scale of Historical Datasets
(12/20/90,  available  from  EPA EMSL-LV),  provides   guidance  that  may  be   used  to  qualify
data  for  inclusion  into   current  databases.    This   EMSL-LV   guidance  stipulates  that  at   least
some   form of  QA/QC  must   be  associated  with  the  historical  data  for  evaluation.     This
QA/QC  may  be  in the  form  of  various types  of  blanks   (method,  field,  etc.),  replicates   field,
analytical,  etc.),  spikes  (matrix,   surrogate   internal   standard,  etc.),  and  PE  samples   (certified
reference   materials,   QC  check   samples   etc.).     A   scoring  mechanism   is   applied   to  these
QA/QC data, and the usability of the sample data is based on the resulting score.
                                                                                      Draft, July 1996

-------
                                                                           	Chapter 2

                                                        Checklist of Laboratory Data Required
                                         to Support Compliance Monitoring for Trace Metals
                                                Determined in Accordance with Method 1669
                                                         and the 1600 Series Analysis Methods
        The  items  listed below describe the  minimum  data  elements  necessary  to validate  trace
metals data collected using the Method for Sampling Ambient Water for Determination of Thrace Metals at EPA
Water Quality  Criteria Levels  (Method  1669)  and  the 1600  Series Analysis Methods.  It  should be
noted  that   since  different  instrumentation  yields  different   data  output,   the  specific  form  of
the data will vary according to the analytical method.

1.       Method Number

        The  method  number of  the   EPA  analytical  method  used  in  conjunction with  Method
1669  must   be  provided.    This  information will  allow  a  data   reviewer  and  user   to  become
familiar  with  the  method,   if necessary,  prior  to  reviewing  the   data.    It  will  also  assist the
reviewer and  user  in  making any  necessary determinations   of the  comparability  of these  data
with  previously  reported  data,  including  qualified  historical   data.    If  more than  one  method
is  needed  to  cover  a  complement  of analytes,  then  all  method  numbers must be  provided.
A  clear  delineation of  the   specific  method  used  for  each  given  analyte is  required.   Also, the
revision  date   or  revision  level and  number/letter  of  the  method  must be given,  so  that the
reviewer or   user  tests  the  results   submitted   against  the  specific  method  used.     Table  1
provides a  list  of  EPA  methods  for  analysis  of  trace metals  along  with  the corresponding
water quality criterion published by EPA for each metal or metal species.

        In  recognition  of  advances  that are  occurring  in analytical  technology,  the  1600  Series
Analysis  Methods  are   performance-based.    That  is,  an  alternate   procedure or  technique  may
be  used  if  the  performance  requirements   in  the reference method(s)   are  met.    The  analyst
must  start  with  one  of  the  methods  as   a  reference,  and may   improve  upon  this  reference
method  to   reduce  interferences  or   lower  costs  of  measurements.    Examples  include   using
alternate  chelating   or   ion  exchange  resins,   alternate  matrix   modifiers,  additional  cleanup
techniques,  or  more  sensitive  detectors.     The  objective  of  allowing  method  modifications  is
to  improve   method  performance  on  the  sample  being analyzed.    At no  time  are changes  that
degrade  method  performance  allowed.    Each  method  details   the  tests  and  documentation
that are required to support equivalent performance.
2.      Detailed Narrative

        A  detailed   narrative   discussing  any   problems   with   the  analysis,   corrective   actions
taken,   and  changes  made  to  the  reference   method  must  be  included  in  a  complete   data
reporting  package.    Reasons  for  changes  to  the  reference  method,  supporting  logic  behind
the technical  approach  to  the  change,  and  the  result  of  the  change  must be  included  in the
narrative.    The  narrative  should  be  written   by  an  analytical  chemist   in  terms  that  another
Draft, July 1996

-------
analytical  chemist  can understand.   The results  of the  review must  be written so  that  the  data
user  can  understand the  reason(s)  for  acceptance/rejection  of  the  data  or  any  changes  to the
reference method.
3.      Data Reporting Forms

        The  complete  data  reporting  package  must  include  data  reporting  forms  that  list  all
samples   analyzed,  the  metals  and  metal  species   determined,  and  the   concentrations  found.
Analytes  detected  in  field samples  at  concentrations  below  the  minimum  level  (ML)   must  be
reported  as  non-detect.    However,  all analyte concentrations  detected  in  blank  samples  must  be
reported,  regardless  of  the  level.    Results  must be listed  for  each sample  analyzed,  including
any dilutions and reanalyses. Metals should be listed by name and CAS Registry number.

        The  ML  is  the quantitation level  as defined  by  the  EPA 1600  series  method  used  for
sample  analysis.    The  laboratory  is  required  to  determine   the   MDL  for  each  analyte  in
accordance  with  the  procedures  described  in 40  CFK  Part  136,  Appendix  B-  Definition   and
Procedure for Determination of Method Detection Limit - Revision 1.11.  That MDL multiplied by 3.18 must be
less than or equal to the ML given in the EPA 1600 Series Analysis Method.

        The   use   of  data  qualifiers   or  flags   by  the  laboratory  is   discouraged.     Rather,
laboratories should  attempt  to  correct  all  analytical  problems   and  provide  a  detailed  narrative
that thoroughly  describes  those  problems  and the  corrective  actions  taken  (see  item  2  above).
Flags or  data  qualifiers  should  be  assigned  by  data users  to  reflect  their specific  data quality
objectives  and  uses  of  the  data.    If the laboratory  submits  data with  internally generated  flags,
the laboratory must provide an explanation of the meaning of the flags used.
4.      Summary of Quality Control Results

        Results  for  all  quality  control  analyses  required  by  the  reference  EPA  method  must
be  presented in  the complete  data  reporting  package.    If more  than  one  method was  used or
if  more   than   one  set   of  samples  was   analyzed,   it  must   be  clearly  evident  which  QC
corresponds to a given method and set of samples.

        Results  for  QC procedures  that  must  be  provided  include, but  are not limited to, the
following (where applicable):

•       Instrument tuning
•       Calibration
•       Calibration verification (initial and following every 10 analytical samples)
•       Initial precision and recovery
•       Ongoing precision and recovery
        Blanks
               Laboratory (method) blanks
               Field blanks
               Calibration blanks
               Equipment blanks
•       Matrix spike/matrix spike duplicates
                                                                                      Draft, July 1996

-------
•       Field duplicates
        Method of standard additions (MSA) results
•       Spectral interference checks
•       Serial dilutions
•       Internal standard recoveries
•       Method detection limits
•       Quality control charts and limits

Table 2 lists the required frequency and purpose of the QC procedures.
5.      Raw Data

        Raw  data  for all  analyses  must  be  kept on file  at  the laboratory (Chapter  2,  Section 7)
and   submitted  for  inspection   to   the   data   reviewer   upon   request.     The  instrument  output
(emission  intensity,  peak  height,  area,  or  other  signal  intensity)   must  be  traceable   from  the
raw  data to  the  final result  reported.   The raw data must be  provided  for  not only  the analysis
of  each   field  sample  but   also  for  all   calibrations,   verifications, blanks,  matrix  spike/matrix
spike  duplicates,  field  duplicates,  and  other  QC   analyses  required   by  the  reference  method.

        Raw  data  are method and instrument  specific  and may  include,  but are  not  limited  to,
the following:

•       Sample numbers and other identifiers
•       Digestion/preparation or extraction dates
•       Analysis dates and times
•       Analysis sequence/run chronology
•       Sample weight or volume
•       Volume prior to each extraction/concentration step
•       Volume after each extraction/concentration step
•       Final volume prior to analysis
•       Injection volume
•       Matrix modifiers
•       Dilution data, differentiating between dilution of a sample or an extract
•       Instrument (make, model, revision, modifications)
•       Sample  introduction    system   (ultrasonic   nebulizer,  hydride   generator,   flow  injection
        system, etc.)
•       Column (manufacturer, length, diameter, chelating or ion exchange resin, etc.)
•       Operating   conditions    (char/ashing   temperatures,   temperature    program,   incident   rf
        power, flow rates, plasma viewing height, etc.)
•       Detector (type, wavelength, slit, analytical mass monitored, etc.)
•       Background correction  scheme
•       Quantitation  reports,  data  system  outputs,  and  other data   to  link the  raw  data to  the
        results reported
•       Direct  instrument  readouts  (e.g.,   strip   charts,  mass   spectra,   printer  tapes,   and  other
        recordings of raw data)  and other data to support the final results
•       Lab   bench  sheets and  copies  of  all  pertinent logbook  pages   for  all  field  and  QC
        sample  preparation and  cleanup   steps,   and  for  all  other  parts  of the  determinations
Draft, July 1996

-------
6.      Example Calculations

        Example  calculations  that  will  allow  an   independent  reviewer  to   determine   how  the
laboratory used  the  raw data  to  arrive  at  a  final  result  must  be  provided in  the data  reporting
package  if   any  adjustments  are  made  to   the  equations   included  in  the  methods.     Useful
examples  include both  detected  and   undetected  compounds.    If  the  laboratory  or   the method
employs  a  standardized  reporting  level   for  undetected  compounds,  this   should  be  made  clear
in   the  example   calculation.      Adjustments   made  for   sample   volume,  dilution,   internal
standardization, etc. should be evident.
7.      Archiving Data on Magnetic Media

        It  is  not   necessary  for  the  laboratory  or  responsible  organization  to  submit  digitized
binary,   hexadecimal,  or  other  raw   signal  recordings   with  the  data   package.    However,  the
laboratory  that  performs  the  analysis  should   archive   these  data  so  that  the  raw reduced  data
can  be  reconstructed,   and   the  laboratory  or  organization  responsible   for   reporting  the  data
should  be  prepared  to  submit  raw  data  on  magnetic   media,  upon  request  by EPA.    Magnetic
media  may  be  required  for  automated  data   review,   for  diagnosis  of  data  reduction  problems,
or for establishment of an analytical database.
8.      Names, Titles, Addresses, and Telephone Numbers of Analysts and QC Officer

        The  names,  titles,  addresses,  and  telephone   numbers  of   the   analysts  who   performed
the  determinations   and  the  quality  control   officer  who  verified  the  results   must  be  included
in  the  data  reporting   package.    If  the  data   package  is  being  submitted  by  a  person  or
organization  other   than   the   analytical  laboratory,  it  is   that  person   or   organization's
responsibility  to  ensure   that   the  laboratory   provides  all  the  data  listed  above  and  that  all
method  requirements  are  met.    For  example,  with  regards  to  effluent  or  ambient  monitoring
data  submitted   by  an  NPDES  permittee  on  a  Discharge  Monitoring  Report  (DMR),  the  task
of collecting and reporting quality control data falls to the permittee.

        In  addition,  the  personnel,  titles,  addresses,  telephone  numbers,  and   name   (if  different
from  the   laboratory  that  analyzed the   field   samples)   of  the  facility  that  cleaned  and  shipped
the  sampling  equipment  and   generated   the   equipment  blanks, the  laboratory  (if different)  that
analyzed  the  equipment  blanks,  and  the  facility  responsible  for   the  collection,  filtration,  and
transport  of  the  field  samples  to the   laboratory  must  be  obtained  and  included  in  the  data
reporting package.
                                                  Table 1
         Method Numbers, Analytical Techniques, Method Detection Limits, and Minimum Levels
1 Method
Technique
Metal
MDL (ug/L)3
ML
Lowest EPA 1
Water Quality 1
Criterion (ug/L)5 |
        Method Detection Limit as determined by 40 CFK Part 136, Appendix B

        Minimum Level (ML) calculated by multiplying laboratory-determined MDL by 3.18 and rounding result to nearest multiple of 1, 2, 5, 10,
        20,50 etc., in accordance with procedures utilized by BAD and described in the EPADraft National Guidance for the Permitting, Monitoring,
        and Enforcement of Water Quality-Based Effluent Limitations Set Below Analytical Detection/Quantitation Levels, March 22, 1994.

        Lowest of the freshwater, marine, or human health WQC at 40 CFR Part 131 (57 FR 60848 for human health criteria and 60 FR 22228 for
        aquatic criteria). Hardness-dependent freshwater aquatic life criteria also calculated to reflect a hardness of 25 mg/L CaCO3, and all aquatic
        life criteria, except chronic criteria for Hg and Se, have been adjusted to reflect dissolved levels in accordance with equations provided in 60
        FR 22228. Hardness-dependent dissolved criteria conversion factors for Cd and Pb also calculated at ahardness of 25 mg/L per 60FR 22228.
        A complete listing of all WQC, including total, dissolved, and levels calculated with a hardness of 25 mg/L CaCO3 is provided in Appendix
        A.

-------
1631
1632
1636
1637
1638
1639
1640
Oxidation/Purge
& Trap/CVAFS
Hydride AA
Ion
Chromatography
CC/STGFAA
ICP/MS
STGFAA
CC/ICP/MS
Mercury
Arsenic
Hexavalent
Chromium
Cadmium
Lead
Antimony
Cadmium
Copper
Lead
Nickel
Selenium
Silver
Thallium
Zinc
Antimony
Cadmium
Trivalent
Chromium
Nickel
Selenium
Zinc
Cadmium
Copper
Lead
Nickel
0.0002
0.003
0.23
0.0075
0.036
0.0097
0.013
0.087
0.015
0.33
0.45
0.029
0.0079
0.14
1.9
0.023
0.10
0.65
0.83
0.14
0.0024
0.024
0.0081
0.029
0.0005
0.01
0.5
0.02
0.1
0.02
0.1
0.2
0.05
1
1
0.1
0.02
0.5
5
0.05
0.2
2
2
0.5
0.01
0.1
0.02
0.1
0.012
0.018
10
0.37
0.54
14
0.37
2.4
0.54
8.2
5
0.32
1.7
32
14
0.37
57
8.2
5
32
0.37
2.4
0.54
8.2
Draft, July 1996

-------
Table 2 ~ Quality Control Requirements, Frequency, and Purpose
Required QC Test
Instrument tuning
Calibration (CAL)
Calibration verification (VKR)
Initial precision and recovery
(IPR)
Method detection limit (MDL)
and minimum level (ML)
Ongoing precision and recovery
(OPR)
Blanks
Matrix spike/matrix spike
duplicate (MS /MSB)
Field duplicate
Method of standard addition
(MSA)
Internal standardization
Spectral interference check
Serial dilutions
Frequency
Prior to calibration
Prior to sample analysis and whenever calibration cannot be verified
Immediately prior to and foil owing the analysis of every batch of 10 or fewer
analytical samples analyzed at the same time
Prior to using the method for the first time and each time a modification to the
method is made
Prior to using the method and whenever there is change that will affect the
MDL and ML
Each sample batch (Sample batch size is method specific. Where not specified,
batch size is 10.)
Equipment blank—Prior to use of any sampling equipment at a given site
Calibration blank— Immediately following each calibration verification
Laboratory (method) blank (BLK)— One method blank per sample batch
Field blank (FBK)— Every ten samples collected at a given site or at least one per
sample site, whichever is more frequent
Each batch of 10 or fewer samples from the same site
Each batch of 10 or fewer samples from the same site
As needed to assure the reliability of results for GFAA or ICP analyses
All analyses by ICP/MS (EPA Methods 200.8 and 200.10)
Prior to using the method for the first time and periodically thereafter as
indicated by instrument stability, type of samples analyzed, and expected
interferences encountered
When analyte concentration is sufficiently high (minimally a factor of 10X the
MDL after dilution)
Purpose
To assure that the instrument will produce results equivalent to instruments in
other laboratories
To establish the working range of the analytical instrument
To verify the average response or curve from the initial calibration
To establish the ability of the laboratory to generate acceptable precision and
recovery
To determine the lowest level at which the analyte can be detected with 99%
confidence that the concentration is greater than zero
To assure that the laboratory remains in control
To assure that contamination of sampling devices and apparatus for sample
collection will be detected prior to shipment to the field site
To assure that contamination of the analytical system will be detected, if present
To assure that contamination of the analytical process will be detected, if present
To assure that contamination of field samples will be detected, if present
To determine bias caused by sample matrix effects
To measure the precision associated with sample collection, preservation,
transportation, and storage procedures, as well as with analytical procedures
To compensate for a sample constituent that enhances or depresses the analyte
signal
To correct instrument drift and other variations in the analytical process
To establish corrections for known interelement spectral interferences
To determine if a chemical or physical interference effect is present

-------
                                                                           	Chapter 3

                                                                  Guidance for Reviewing Data
                                                      from the Analysis of Trace Metals Using
                                          Method 1669 and the 1600 Series Analysis Methods
        Use   of  the   guidelines  provided   below,   or   of  similarly   developed   standardized
protocols,  is  highly   recommended   as  a   tool  with   which   Regional   and  State  permitting
authorities   can  standardize   data   inspection   and   acceptance   procedures   and  minimize
differences  that  might   otherwise  result  between  data  reviewers  and/or  permittees  responsible
for submitting data.   A Data Inspection Checklist has also been developed and is  provided in  the
following  chapter.     This  checklist   provides  a  standardized   format  for  documenting   the
findings   of  each  data  inspection  and  an   additional  tool   for  standardizing  the   data review
process within a regulatory agency.
1.       Purity and Traceability of Reference Standards

        The  accuracy   of  any   non-absolute  empirical   measurement   is   dependent   on   the
reference  for that measurement.    In  determining pollutants  in  water  or other  sample  matrices,
the  analytical  instrument  and  analytical  process   must  be   calibrated  with  a  known  reference
material  of  documented purity  and  traceability.    This  information  need not  be  provided with
every sample analysis.    Rather,  it should  be  maintained  on  file at  the  laboratory  and  provided
upon  request.     When  analyses  are  conducted  in  a  contract  laboratory,  such  documentation
should  be provided  to  the permittee  the  first time that the  laboratory  is employed  for  specific
analyses and updated as needed.
2.      Number of Calibration Points

        The  1600  Series  Analysis  Methods  specify  that  a  minimum  of  three  concentrations  are
to be  used when  calibrating  the  instrument.   One  of these  points must be  the Minimum  Level
(ML,   see   item   5),  and  another must  be  near   the   upper  end  of  the   calibration  range.
Calibration  must   be  performed  for   each  target  metal  before  any   samples  or  blanks  are
analyzed.    The use  of the  ML  as  a  point  on  the  calibration  curve is  the  principal  means by
which to assure that measurements made at this quantitation level are reliable.

        The  data  reviewer  should  review  the  points used  by  the  laboratory to  calibrate  the
instrument  and make  certain  that the  calibration  range  encompasses  the  Minimum  Level  and
that   all  sample  and  QC   measurements   are   within   the   calibration  range.     Samples   that
produced    results  which   exceeded  the   calibration  range   should   have   been  diluted   and
reanalyzed   in  accordance   with   the  specifications  detailed  in the  1600  Series  Analysis  Method
that  was  used by  the laboratory.   The diluted sample  results need only apply  to  those analytes
that  exceeded  the  calibration  range of  the  instrument.   In  other words,  it is  acceptable to  use
data for  different  analytes  from  different  levels   within  the   same  sample.   Some  flexibility  may
Draft, July 1996

-------
be  exercised  in  acceptance  of  data  that  are  only  slightly above  (<10%)  the  calibration  range.
Such data are generally acceptable as calculated.

        If data  from  an  analysis  of  the  diluted  sample  are  not  provided, limited use  can  be
made  of the  data that  are  above  the  calibration  range  (>10%).   The response  of  the  analytical
instrument  to  concentrations of analytes  will  eventually  level  off  at concentrations  above  the
calibration  range.    While  it  is  not  possible  to  specify  the  concentration  at  which  this  will
occur,  it  is   generally  safe  to  assume  that  the  reported  concentration  above   the  calibrated
range  is  a lower  limit  of  the  actual  concentration.   Therefore,  if the  concentration  above  the
calibration  range  is   also   above  a  regulatory   limit,  it  is  a  virtual  certainty  that  the  actual
concentration would also be above that limit.
3.      Linearity of Calibration

        The    relationship   between   the   response    of   an   analytical   instrument    to    the
concentration  or  amount  of an  analyte  introduced  into  the  instrument is  referred  to  as  the
"calibration  curve".    An analytical  instrument can  be  said  to  be  calibrated  in  any instance  in
which  an  instrumental  response  can  be  related  to  the  concentration  of  an   analyte.    The
response  factor  (RF,  calculated  for external  standard  calibration)    or  relative  response  factor
(RRF,  calculated   for  internal   standard   calibration)   is   the   ratio   of  the   response  of  the
instrument  to   the   concentration  of the  analyte  introduced  into  the  instrument.    Equations
for calculating RFs and RRFs are provided in the 1600 Series Analysis Methods.

        While  the  shape of calibration  curves can  be  modeled by  quadratic  equations  or  higher
order  mathematical   functions,  most  analytical methods  focus  on   a   calibration   range  in  which
the  linear   calibration   is  essentially  a   function  of  the  concentration  of   the   analyte.     The
advantage  of  the   linear calibration  is  that  the  RF  or  RRF represents  the  slope  of  calibration
curve,  simplifying  calculations   and  data   interpretation.     The   1600   Series  Analysis  Methods
contain  specific   criteria   for   determining  the   linearity   of  calibration   curves   determined  by
either  an  internal  or  external   standard  technique.    When  the  applicable  criterion is   met,  the
calibration curve is  sufficiently   linear to  permit  the  laboratory to  use  an  average RF  or  RRF,
and it  is  assumed that  the  calibration  curve  is  a straight  line that passes  through  the  zero/zero
calibration point.    Linearity  is  determined  by  calculating  the  relative  standard   deviation  (RSD)
of  the  RF  or  RRF  for  each  analyte  and  comparing  this   RSD  to  the   specified limit.    The
specific  acceptance  criteria   are  listed  in  the  Data  Inspection  Checklist  (Chapter  4,  Item  12)
and in  the   1600  Series  Analysis Methods.    These  methods  also  include  alternative  procedures
to be used in the event the linearity criteria fail specifications.

        The  laboratory  must provide the  RSD   results  by  which  an  independent  reviewer  can
judge  linearity, even  in  instances in  which  the laboratory is  using  a  calibration curve.   In  these
instances,  the  data  reviewer should  review  each  calibration  point to  assure  that  the  response
increases  as   the   concentration   increases.     If  it  does  not,  the  instrument  is  not   operating
properly, and the data should not be considered valid.
10                                                                                     Draft, July 1996

-------
4.      Calibration Verification

        Calibration  verification  involves  the  analysis   of  a   single  standard,   typically   in   the
middle  of  the  calibration  range,  at  the  beginning  (and,  in  some  cases,  at  the  end)  of  each
analytical   shift.     The  concentration  of  each   analyte  in  a  reference  standard  is   determined
using  the  initial  calibration  data  and  compared  to  specifications  in  the method.   If  results  are
within  the   specifications,  the  laboratory  may  proceed  with  analysis  without recalibrating.    The
initial   calibration   data  are  then  used   to   quantify  sample  results.     Specific  criteria   for
acceptance   of   calibration   verifications   are   provided   in   the   Data   Inspection   Checklist
(Chapter 4, Item 17) and the 1600 Series Analysis Methods.

        Calibration  verification, which  is  used  in  the  1600  Series  Analysis  Methods,  differs  in
concept  and  practice  from  "continuing  calibration",   which   is  used   in   the   SW-846  methods
and   in   the  Superfund   Contract  Laboratory   Program  (CLP).     In  continuing  calibration,   a
standard  is  analyzed  and  new  response  factors are  calculated  on  the  basis   of  that  analysis.
If the  new factors are  close to  the  average  from the  initial calibration,  all subsequent sample
analyses   are  conducted   using  the   new   response  factors.     The   degree  of  "closeness"  is
generally  measured as  the  percent difference  between  the  old and  new  factors.    The  problem
with  continuing  calibration  is that  it  amounts   to  a  daily  single-point calibration.   Information
about  the  behavior  of the   instrument  at concentrations  above  and below  this  single  standard
can only be inferred from the initial multiple-point calibration.

        The   1600  Series   Analysis   Methods   require   calibration   verification   after   every   ten
samples.     Calibration  verification  is  performed   by  analyzing   an  aliquot  of  the  mid-point
calibration  standard,   and   obtaining  results  that  meet   the  specifications  contained   in   the
methods.    These   specifications  are  given  for  each method  and  metal as  a percentage  of  the
recovery  of the  mid-point  calibration standard.    If any individual  value  falls  outside  the  range
given,  system  performance   is   considered   unacceptable,   and   the  laboratory   may  either
recalibrate  the  instrument  or  prepare  a  new  calibration  standard   and make a  second  attempt
to  verify  calibration.    The  data  reviewer  should verify   that  each   batch of  10  samples  is
associated with a calibration verification that meets the required performance criteria.
5.      Method Detection Limit and Minimum Level

        The  Minimum  Level  (ML)   is  defined  in  the  1600   Series  Analysis   Methods  as  the
lowest  level  at  which  the  entire  analytical system  gives  a  recognizable  signal  and  acceptable
calibration  point.    Therefore,  each  1600   Series   Analysis  Method   requires  that  the  calibration
line or curve for each analyte encompass the method-specified ML.

        The  1600  Series  Analysis  Methods  also  require  each  laboratory  to  perform  a  method
detection  limit  (MDL)  study for  each  analyte  in  accordance  with   the  procedures  given  in  40
CFR  Part   136,   Appendix   B.     The  MDL  studies  are   conducted  to   demonstrate  that  the
laboratory can  achieve  the  MDLs  listed  in  the  methods.   MDL  determinations  must  be  made
the  first  time  that the  laboratory  utilizes   the  method  and  each  time  the  laboratory  utilizes  a
new instrument or modifies the method in any way.

        Each  MDL and ML listed  in the  1600   Series  Analysis  Methods  represents  the  results
of  MDL  studies   conducted  by  the  EPA's Engineering  and  Analysis   Division  as  part  of  its
Draft, July 1996                                                                                     11

-------
effort  to validate  the  methods.   The  MDL  studies were conducted  by  at  least  one  laboratory
for  each method  and  metal  in  accordance  with  the  procedure  given   in  40  CFR  Part  136,
Appendix  B.    The  MLs   shown  in  Table   1  were  calculated   by  multiplying  each  laboratory-
determined  MDL  by 3.18  and  rounding  the  result  to  the  nearest multiple of 1,  2,  5,  10,  20,
50,  etc.  in   accordance  with  the  procedures  described  in  the  EPA  Draft  National  Guidance for  the
Permitting, Monitoring,  and Enforcement  of Water Quality-Eased Effluent  Limitations  Set  Below Analytical
Detection/Quantitation Levels, March 22, 1994.

        The   1600  Series   Analysis   Methods  and  Chapter   2   of  this   document   require  the
laboratory  to report  the concentration  of all  sample  results  that  are  at  or  above  the ML.   It
should  be   noted  that  this  ML  is  a  sample-specific  ML  and,  therefore,  reflects   any  sample
dilutions  that  were  performed.     If  sample  results   are  reported  below  the  ML,  the   data
reviewer  should  require  the  responsible   party  to  correct  and  resubmit  the  data,   or  if  this
course  of  action  is  not  possible,  the  reviewer  should  determine  the   sample-specific ML  and
consider results below that level to be non-detects for regulatory  purposes.

        If  sample  results  are  reported above  the  ML,  but  are  below  the   facility's  regulatory
compliance   level,   then the  data   reviewer   should  consider   the  results  to   suggest  that  the
pollutant  has been  detected but  is  compliant with  the  facility's  permit  (assuming  that  all  QC
criteria  are   met).    If  sample  results  are  reported  above  the   regulatory compliance  level,  the
data  reviewer  must  evaluate  laboratory   QC  samples  in  order  to  verify   that   the level   of
pollutant  is  not  attributable to  analytical bias.    In  addition,  the  data  reviewer  must evaluate
all  blank  sample  results  in  order  to  determine  if  the level  of  pollutant  detected  may   be
attributable to contamination.

        Although  sample  results  are  to  be   reported  only  if  they  exceed  the  ML,  all  blank
results  are   to  be  reported,  regardless of  the  level.    This  reporting   requirement  allows  data
reviewers the  opportunity to  assess   the   impact   of any  blank  contamination  on  sample  results
that are reported above the ML.

        It  is  important  to  remember  that if  a  change  that will  affect  the  MDL  is  made  to  a
method,  the  MDL  procedure  must  be repeated  using the  modified  procedure.    Changes  may
include   alternate   digestion,   concentration,  and   cleanup   procedures,   and    changes   in
instrumentation.     Alternate   determinative  techniques,  such   as  the   substitution   of  a
colorimetric  technique  or  changes  that  degrade  method  performance  are  not allowed.    The
data  reviewer  should  verify   that  method   modifications were  appropriate   and   were  capable
of producing the desired MDLs.

        The  procedures  given  in  this  document   are  for  evaluation of  results  for determination
of  regulatory   compliance,   and  not   for  assessment   of  trends,  for  triggering,   or  for  other
purposes.    For   such   other  purposes,  the  reporting  of all   results,   whether  negative,  zero,
below the MDL, above the  MDL but below the ML,  or above the  ML, may be  of  value  and
may  be  required   by  the  permitting   authority  as  necessary   to   enforce  in   a  particular
circumstance.   Dealing  with  the  multiplicity  of  consequences  presented   by   such   results,  either
singly or in combination, is beyond the present scope of this  document.

6.      Initial Precision and Recovery
12                                                                                   Draft, July 1996

-------
        The  laboratory  is  required   to  demonstrate  its  ability  to  generate  acceptable  precision
and  accuracy  data  using  the techniques  specified  in  the  1600  Series  Analysis  Methods.   This
test,  which  is  sometimes   termed  the  "start-up  test",  must  be  performed  by  the  laboratory
prior  to  the  analysis  of  field  samples  with  the  specified  methods   and  prior  to  the   use   of
modified  versions  of  the  methods  on   field  samples.     EPA's   experience   has  been  that
laboratories   that  have  difficulty  passing  the  start-up  test  have  such  marginal performance  that
they will have difficulty in the routine practice of the method.

        The  test  consists  of spiking  four  aliquots  of  reagent  water  with  the metals  of  interest
at 2  -  3  times  the  ML  concentrations listed  in the  methods and analyzing these  four  aliquots.
The  mean   concentration   (x)  and   the   standard  deviation   (s)   are   then  calculated  for  each
analyte  and  compared to  the  specifications  in  the  methods.    If the  mean  and  the  standard
deviation are  within the limits, the laboratory can use the method to analyze field samples.

        If the  start-up  test data  fail  to meet the specifications in  the  method,  none of the data
produced  by the laboratory  can  be  considered to  be  valid.   If the  laboratory  did  not  perform
the start-up  tests, the  data cannot be  valid,  unless  all  other  QC criteria have  been  met  and  the
laboratory  has  submitted  IPR  (and  associated  instrument  QC)  data   that  were  generated  after-
the-fact by   the  same  analyst on  the  same  instrument.   If  these  conditions  are  met,  then  the
data  reviewer   may   consider  the  data  to  be  acceptable  for most  purposes.    NOTE:   The
inclusion  of this  alternative  should  not  in  any  way  be construed  to  sanction  the  practice   of
performing   IPR  analyses  after   the  analysis   of  field  samples.     Rather,  EPA   believes  that
demonstration   of   laboratory   capability  prior   to  sample  analysis   is   an   essential  QC
component;   this  alternative  is  provided  only  as  a  tool to permitting  authorities  when data
have  already been  collected  without  the  required  IPR  samples.    Once   the  problem has  been
identified,  all   responsible  parties  are  expected  to   implement  corrective  action   necessary   to
ensure that it is not repeated.

        It is  important to  remember that  if a  change  is  made  to  a  method,  the  IPR procedure
must  be  repeated  using  the modified  procedure.    If  the  start-up   test  is  not  repeated  when
these  steps  are  modified  or added, any  data  produced by  the  modified methods  cannot  be
considered to be valid.
7.      Analysis of Blanks

        Because  trace  metals  are  ubiquitous  in  the  environment,  the  precautions  necessary  to
preclude  contamination   are   more   extensive   than   those  required  to  preclude   contamination
when   synthetic  organic   compounds   and   other  non-ubiquitous   substances  are   determined.
EPA has   found  that  the  greatest  potential  for contamination  of  samples  analyzed for  trace
metals  has  been  from  atmospheric  input  in  the   field  and  laboratory  and  from  inadequate
cleaning  of  sample  bottles   and  labware.     In  order  to  prove   that  such  contamination  is
avoided  during  sampling,  sample  transit,   and  analysis,  Method  1669  and   the   1600  Series
Analysis  Methods  specify  the   collection   and  analysis  of  numerous   blank  samples.    These
include:

•       Equipment blanks  that are  collected prior to  the  use   of any  sampling  equipment  at a
        given   site  and  provide   a means  for  detecting  contamination  of  sampling  devices  and
        apparatus prior to shipment to the field site,


Draft, July 1996                                                                                   13

-------
•       Field  blanks  that  are collected for  each  batch of 10  or  fewer  samples  from  the same
        site and provide a means of detecting contamination that arises in the field,

•       Calibration   blanks   that  are   analyzed   immediately   after   each  calibration   verification
        and  provide  a   means  of  detecting  contamination  that   arises  from the  analytical   system,
        and

•       Laboratory  (method)  blanks  that   are  analyzed  for  each  batch  of samples  analyzed   on
        a  particular  instrument  and  provide  a   means   of  detecting   contamination  from  the
        analytical process.

        While  the  analysis  of  a  minimum  of four   blank  samples per  site  may seem  to   be
excessive,  particularly  when  very  few  (e.g.,  <  5)  samples  are  collected,  EPA  has  found that
the validity  of entire  studies  may be  suspect  when pollutants  are  identified  in samples  that  are
not associated with  each of these  blanks.   In  general,  it  is  not  necessary for a facility to  report
the  results  of  equipment  blank  analyses  unless  contamination  is  identified  in  field   blanks.
Therefore,  the  permittee   should   obtain  equipment   blank   results  from  its  cleaning   facility,
maintain  these  results   on  file,  and  provide  them  to  the  permitting  authority  upon  request.
The data reviewer  should  evaluate  equipment  blank  results  only  if it  is necessary   to  identify
potential sources of contamination present in field blanks.

        Controlling   laboratory  contamination  is  an  important   aspect   of  the  quality  assurance
plan  for   the   equipment-cleaning   facility,  laboratory,  and  field  team.     Each   party   should
maintain  records  regarding  blank  contamination.     Typically,   these records   take  the  form  of
a  paper  trail  for  each  piece   of equipment  and control  charts,  and  they  should  be  used  to
prompt  corrective  action   by  the  party   associated  with  the  contamination.     For   example,   if
records   at  a  single   site  suggest  that   equipment  blanks,  laboratory  blanks,   and  calibration
blanks  are  consistently  clean   but  that   field  blanks   show  consistent  levels  of  contamination,
then  the   field   sampling   team  should   re-evaluate  their  sample  handling  procedures,  identify
the  problem,  and  institute corrective actions  before  collecting   additional   samples.    Similarly,
equipment  cleaning  facilities  and  laboratories  should   utilize  the   results  of  blank   analyses  to
identify and correct problems in their processes.

        Unfortunately,  it is  often  too  late  for  corrective  action  if data  are received  that   suggest
the  presence  of  uncontrolled  contamination  that  adversely  affects the   associated  data.    The
exception  to  this   rule  is the   case  in  which  the  field  and  equipment  blanks  show   no
discernable   levels   of  contamination,   contamination   is   detected   in   the   laboratory   or
calibration  blanks,   sample   holding  times  have   not  expired,   and   sufficient  sample  volume
remains   to  allow  the  laboratory  to   identify  and  eliminate  the   source  of  contamination and
reanalyze   the  associated  sample(s).    In  all  other  cases,  the   reviewer  must  exercise   one  of
several options listed below when making use of the data.
14                                                                                    Draft, July 1996

-------
•       If a contaminant  is  present  in  a blank  but is  not present in  a sample, then  there  is little
        need  for  concern  about  the sample result.   (It  may be  useful,  however,  to occasionally
        review the  raw  data  for  samples  without the  contaminant  to  ensure that the  laboratory
        did not edit the results for this compound.)

•       If the sample contains the  contaminant  at levels  of at least  10  times that in  the  blank,
        then  the  likely  contribution  to  the sample   from  the  contaminant  in  the sample  is  at
        most  10%.    Since  most  of  the  methods in  question are  no  more  accurate  than  that
        level,  the  possible   contamination  is  negligible,  and  the  data  can  be  considered   to  be
        of acceptable quality.

•       If the sample  contains  the contaminant  at  levels  of at  least  5  times  but less than  10
        times   the   blank  result,  the  numerical  result  in  the   sample  should  be  considered  an
        upper  limit  of the  true  concentration,  and  data  users  should be  cautioned  when  using
        such data for enforcement purposes.

•       If the sample  contains  the  contaminant  at levels below  5  times  the level in  the  blank,
        the sample   data  are  suspect  unless  there are  sufficient  data  from  analyses  of  multiple
        blanks to  perform   a   statistical  analysis  proving  the  significance  of  the  analytical  result.
        Such statistical analyses are beyond the scope of this guidance.

•       If blank  contamination  is   found  in  some  types  of QC  samples  but  not   others  (e.g.,
        only in the  laboratory  blank but not in  the  field  blank), the  data user  should  apply  the
        guidelines  listed above, but may   also  use   this  information  to  identify  the  source  of
        contamination and take corrective  actions to prevent future recurrences.

        There  are   two  difficulties   in  evaluating   sample   results   relative  to  blank contamination.
First, the  reviewer  must  be  able  to associate  the samples with the  correct  blanks.   Field  blanks
are  associated  with  each  group  of   field  samples   collected  from  the  same  site.     Calibration
blanks  are associated  with  samples by  the date  and  time  of  analysis  on  a specific  instrument.
Laboratory  (method)  blanks   are   associated  with  each  batch  of   10  samples  prepared  and
digested in accordance with a  particular method  during  a  single  shift.   If the  reviewer  cannot
associate  a batch   of samples  with  a  given  blank,  the   reviewer  should  request  this  association
from the laboratory so that the results for the samples can be validated.

        The  second  difficulty  involves   samples   that  have  been  diluted.    The  dilution   of  the
sample   with   reagent  water  represents  an  additional   potential   source  of  contamination  that
will  not  be   reflected  in   the results  for the   blank  unless  the  blank  was  similarly   diluted.
Therefore,  in  applying  the   10-times   rule  stated above,  the  concentration  of  the   sample  is
compared  to  the  blank results multiplied   by  the  dilution  factor  of  the  sample.   For instance,
if 1.2  ppb of a contaminant  is  found  in the  blank,  and  the associated sample  was  diluted  by
a  factor  of six  relative  to  the  extract  from  the   blank  prior to  analysis, then  the diluted  sample
result would have  to be  greater than  1.2 x 6  x  10  or 72 ppb  to be  acceptable.   Diluted  sample
results   between   36   and  72  ppb   would   be  considered   an  upper  limit   of  the   actual
concentration,  and   diluted   sample  results  that  were  less   than  36   ppb  would  be   considered
unacceptable  in  the absence  of  sufficient  blank data   to   statistically  prove  the  significance  of
the result.
Draft, July 1996                                                                                    15

-------
        In  most  cases,   the  practice   of  subtracting  the   concentration  reported  in  the  blank
from   the   concentration  in  the   sample  is  not  recommended   as   a   tool  to  evaluate   sample
results  associated with  blank  data.    One  of the  most common  problems with  this  approach
is   that   blank  concentrations  are   sometimes   higher   than  one   or   more  associated   sample
results, yielding negative results.

        Nearly all  of  the  1600  Series  Analysis  Methods  are  capable  of producing MDLs  that
are at  least  10  times  lower  than  the  lowest  water  quality  criteria  (WQC)   published   in  the
National  Toxics  Rule.     Since   most  discharge  permits  require monitoring  at levels  that  are
comparable  to  or  higher  than  the  WQC  published  in  the  National Toxics Rule,  EPA  believes
that,  in  nearly  all  cases, laboratories  should  be  capable  of  producing  blank  data  that   are  at
least  10  times  less   than  the   regulatory  compliance  level.    It   should  also be  noted  that
laboratories  cannot  be  held  accountable  for  contamination  that is   present  in   field  blanks  but
not  present  in  laboratory  blanks;  in  such   cases  the  sampling  crew  should  take   corrective
measures   to   eliminate   the  source   of  contamination   during  their  sample  collection  and
handling steps.
8.      Ongoing Precision and Recovery

        The   1600  Series   Analysis  Methods   require  laboratories  to  prepare  and   analyze   an
"ongoing  precision  and  recovery"  (OPR)  sample  with each  batch  of up to  10  samples  started
through  the  extraction  process  on the  same  twelve  hour shift.   This  OPR sample  is  identical
to the  aliquots  used  in  the IPR  analyses  (see  Item  6), and  the results of the OPR  are used  to
ensure  that   laboratory  performance   is  in  control  during the   analysis  of  the  associated  batch
of field samples.

        The  data  reviewer  must  verify  that the  OPR sample  has  been  run  with each  sample
batch and that  the  applicable recovery criteria in  the  analytical  method  have  been met.    If  the
recovery  criteria  have  not  been met,  the  reviewer  may  use  the  following  guidelines  when
making use of the data:

•       If the concentration of the  OPR is above method  specifications   but  that  analyte  is  not
        detected in  an  associated  sample, then  it  unlikely that  the  sample  result  is   affected  by
        the failure in the OPR.

•       If the  concentration  of  the   OPR  is  above  method   specifications   and  that  analyte  is
        detected in  the  sample,  then  the numerical  sample  result may  represent  an  upper limit
        of the  true  concentration, and data users  should be  cautioned when  using the  data  for
        enforcement purposes.

•       If the  concentration  of  the  OPR is  below  method   specification   but  that  analyte  is
        detected in  an  associated  sample, then  the  sample  result  may represent the  lower limit
        of the true  concentration for that analyte.

•       If the concentration of the  OPR is below  method  specification  and  that  analyte  is  not
        detected  in  an  associated  sample,  then  the  sample  data  are   suspect  and  cannot   be
        considered valid for regulatory compliance purposes.
16                                                                                   Draft, July 1996

-------
        If  the  OPR  standard  has  not  been  run,  there  is  no  way to  verify  that  the laboratory
processes were  in  control.    In  such  cases, a  data  reviewer  may be  able  to  utilize  the  field
sample  data by examining  the  matrix  spike  recovery  results  (see  item  9), the  IPR  results,  OPR
results  from  previous   and  subsequent  batches,  and  any   available  historical  data  from  both
the  laboratory  and  the  sample  site.    If  the   matrix spike  results  associated  with  the  sample
batch  do  not  meet  the performance  criteria in  the  methods,  then  the  results  for that  set  of
samples  cannot be   considered   valid.    If  the   laboratory's  IPR  results  and  the  matrix  spike
results  associated  with  the  sample   batch   in   question  meet  the  all   applicable   performance
criteria  in  the  methods,  then  the  data  reviewer  may  be   reasonably  confident  that laboratory
performance was  in  control  during  field   sample  analysis.    This  level  of  confidence may   be
further  increased if  there is  a  strong  history of both  laboratory  performance with the  method
and  method  performance with  the  sample  matrix  in  question,  as indicated  by additional  OPR
and matrix spike data collected from the laboratory and samples from the same site.
9.      Precision and Recovery of Matrix Spike and Matrix Spike Duplicate Compounds

        The   1600  Series   Analysis   Methods  require   that   laboratories   spike   the  analytes   of
interest  into   duplicate  aliquots  of   at  least  one  sample   from  each   group  of   ten  samples
collected  from  a single  site.    The first  of these  spiked  sample  aliquots  is known  as the  matrix
spike  sample;  the   second  is  known  as  the  matrix  spike  duplicate.     These  spiked  sample
aliquots  are  used to  determine  if the  method  is  applicable  to  the  sample  matrix  in  question.
The   1600  Series   Analysis   Methods   are   applicable  to   the  determination   of  metals   at
concentrations   typically  found  in  ambient  water   samples   and   certain  treated  effluents  (e.g.,
the  part-per-trillion  to  low part-per-billion  range).    These   methods  may  not  be  applicable   to
marine  samples  and  many  effluent  and  in-process   samples   collected   from  industrial
dischargers.   Therefore, it  is  important  to  evaluate  method performance  in  the  sample  matrix
of interest.

        In evaluating  matrix spike  sample  results,  it is  important  to  examine  both  the  precision
and accuracy   of the  duplicate  analyses. Precision  is  assessed  by  examining  the relative  percent
difference  (RPD)  of  the  concentrations found  in  the  matrix  spike  and  matrix  spike  duplicate
samples,  and   comparing   the   RPD   to   the  acceptance   criteria  specified   in   the  analytical
method.    If  the  RPD  of  a  matrix  spike/matrix spike  duplicate  pair  exceeds  the  applicable
criterion,  then  the  method  cannot  be  considered  to   be applicable   to  the  sample  matrix,  and
none of the associated sample data can be accepted for regulatory compliance purposes.

        If RPD  criteria  are met, the method is considered  to   be capable  of producing precise
data in  these   samples,  and the  data reviewer must  then verify  that  the method  is capable   of
producing  accurate  data.    Accuracy  is  assessed  by  examining  the   recovery  of  compounds   in
the  matrix  spike and  matrix  spike duplicate  samples.    If the recovery  of the matrix  spike  and
duplicate  are   within  the  method-specified  limits,  then  the   method   is  judged  to   be  applicable
to  that sample matrix.   If,  however,  the  recovery  of  the  spike  is not within  the recovery range
specified,  either the method  does  not  work  on the  sample, or the  sample preparation  process
is out of control.

        If the  method is  not appropriate  for the  sample  matrix, then  changes  to  the  method
are  required.    Matrix  spike  results  are  necessary  in   evaluating  the  modified  method.    If  the
Draft, July 1996                                                                                    17

-------
analytical   process   is  out   of  control,  the  laboratory   must  take  immediate  corrective  action
before any more samples are analyzed.

        To    separate   indications   of   method    performance    from    those   of   laboratory
performance,   the  laboratory   should   prepare  and  analyze  calibration   verification   standards
and  OPR  samples.    If  the  results  for either of these  analyses  are  not within  the  specified
range,  then  the analytical  system  or process  must be  corrected.    After the  performance of the
analytical   system  and  processes  have  been  verified  (through  the  successful  analysis  of   CCV
and  OPR  samples),  the  spike   sample  analysis should  be  repeated.    If the  recovery  of the
matrix  spike  and  duplicate  are  within   the   method-specified   range,  then   the  method   and
laboratory   performance  can  be   considered   acceptable.     If,   however,   the  recovery   of  the
matrix  spike  does  not  meet  the  specified  range,  the  laboratory  should  attempt   to  further
isolate the  metal  and repeat  the  test.    If  recovery of  the metal  remains   outside the  acceptance
criteria,  the  data  reviewer  may  apply  the  following guidelines   when  attempting to  make  use
of the data:

•       If the  recovery  of  the  matrix  spike  and   duplicate are  above  method  specifications  but
        that  metal  is  not  detected  in  an  associated  sample  or  is  detected below the  regulatory
        compliance  limit, then it unlikely  that  the  sample result  is  affected by  the failure in the
        matrix spike.

•       If the  recovery  of  the  matrix  spike  and  duplicate are  above method specifications  and
        that  metal  is  detected  in an  associated  sample  above  the  regulatory  compliance   level,
        then  the  sample   result  may  represent   the  upper limit   of the  true   concentration,  and
        the data should not be considered valid for regulatory compliance purposes.

•       If  the    concentration   of  the    matrix    spike   and    duplicate   are    below   method
        specifications  but  that metal  is  detected in  an associated sample,  then the  sample  result
        may represent the  lower  limit  of the  true  concentration for  that  metal.    If the  metal
        was  detected  in  the  sample  at  a  concentration  higher   than the  regulatory   compliance
        limit,  then  it  is  unlikely that  the sample  result  is adversely affected  by  the  matrix.    If,
        however,   the  metal  was   detected   below  the   regulatory  compliance   limit,   the   data
        should not be considered valid for regulatory compliance purposes.
10.      Statements of Data Quality for Spiked Sample Results

        The  1600  Series  Analysis  Methods  specify  that  after  the  analysis  of five  spiked  samples
of  a  given  matrix  type,  a  statement  of  data  quality  is  constructed  for each  analyte.    The
statement  of data quality for each  analyte  is  computed  as  the  mean percent  recovery  plus and
minus   two  times   the   standard   deviation  of  the   percent   recovery  for  the   analyte.     The
statements  of  data  quality  should  then  be  updated  by  the  laboratory  after  each  five  to  ten
subsequent spiked sample analysis.

        The  statement of  data  quality  can  be  used  to  estimate  the true  value  of a  reported
result  and   to   construct  confidence  bounds  around  the  result.    For   example,   if  the  result
reported for  analysis  of  selenium  is  10 ppb,  and  the  statement  of  data quality  for  selenium
is  84%  ±  25%  (i.e.,  the  mean  recovery is 84% and the  standard  deviation  of the  recovery is
18                                                                                   Draft, July 1996

-------
25%),  then  the  true  value  for  selenium  will  be  in  the  range  of  9.4  -  14.4 ppb,  with  95%
confidence.  This range is derived as follows:

        Lower Limit  = [(10 - .84) - (10 x .25)] = [11.9 - 2.5] = 9.4 ppb
        Upper Limit  = [(10 - .84) + (10 x .25)] = [11.9 +  2.5]  = 14.4 ppb

        Many  laboratories  do  not  provide  the  data   quality   statements  with  the  sample  results,
in  which  case   the  data  reviewer  must  determine   if  the  data  quality   statements  are  being
maintained  for  each  analyte  and  may need to  obtain the data.   If necessary,  the  reviewer  can
construct the  data  quality  statement  from the  individual  data  points.    The  lack  of  a  data
quality   statement  does  not  invalidate   results  but   makes  some   compliance   decisions  more
difficult.   If statements  of  data  quality  are  not  being  maintained  by  the  laboratory,  there  may
be  increased  concern  about  both  specific  sample  results  and  the  laboratory's  overall  quality
assurance program.
11.      Statements of Data Quality for Spiked Reagent Water Results

        In  addition   to  statements  of  data  quality  for  results  of  analyses   of  the  compounds
spiked  into   field  samples,  the  1600  Series Analysis  Methods  require  that  statements   of data
quality  be  constructed  from the  initial and ongoing precision and recovery  data.   The  purpose
of  these  statements  is  to   assess  laboratory  performance  in  the practice  of  the  method,  as
compared  to  the   assessment  of method  performance  made  from  the   results   of  spiked  field
samples.   Ideally,  the two  statements  of  data quality would  be  the  same.   Any  difference  could
be attributable to either random error or sample matrix effects.


12.      Field Duplicates

        Method  1669  requires  the  collection of  at least  one  field  duplicate  for  each  batch  of
field  samples  collected from  the same site.   The  field duplicate provides  an  indication of the
overall   precision  associated  with   entire   data   gathering   effort,  including   sample   collection,
preservation,   transportation,   storage,  and  analysis  procedures.     The   data   reviewer   should
examine  field  duplicate  results  and  use  the following  equation  to  calculate   the  relative  percent
difference between the duplicate and its associated samples.
                                       RPD  - 2
                                                    (D1+D2)

        where:
                 Dl = concentration of the analyte in the field sample
                 D2 = concentration of the analyte in the duplicate field sample
        If the  analyte of interest was not  detected in  either  replicate  of  the  field  sample,  then
the  RPD  will  be  zero.    If  the  analyte  was  detected  in  each  field  sample  replicate, but the
results  are   highly   disparate   (indicated  by   a  large   RSD),  the  reviewer   should  apply  the
following guidelines when making use of the data:
Draft, July 1996                                                                                    19

-------
•       If  the   analyte  was   detected  in  each   replicate  and  at   similarly  variable   concentrations
        in  the  blank  samples,  then  the  field  sample  variability  may  be  attributable  to  variable
        contamination, and the data may not be valid for regulatory compliance purposes.

•       If  the   analyte   was   detected   in  each   replicate   at   a  concentration  well  above  the
        regulatory   compliance  level,  but  was   not   detected  in  the  associated  blank  samples,
        then it is likely that the sample results are not adversely affected.

        Ideally,   the  RPD  between  field   duplicates  and   MS/MSD   samples  will  be   identical.
Any  difference   between   the  two is  attributable  to   variability  associated with the  field  sampling
process.
20                                                                                      Draft, July 1996

-------
                                                                        	Chapter 4

                                                                   Data Inspection Checklist
       The   following pages  contain  a  data  inspection  checklist  that  may  be  used  by  data
reviewers,  laboratory   personnel,   and  other   parties   to  document  the  results   of  each  data
inspection in a standardized format.
Draft, July 1996                                                                              21

-------
                                                 Data Inspection Checklist
                                                    Summary Information
   1.  Name of Reviewer:
                      Tide:
                       Required Samples
                                       Sample Results Provided
        Sample Location or
            Sample ID
Analyte(s)
Sample Location or
    Sample ID
Analyte(s)
   2.  Metiiod Used:

   3.  Total No. of analytical shifts per instrument (determined from analysis run 1
            Instrument
   4.  Total No. of CCVs Required:
   (one for each 10 samples after the
   first 10 samples on each instrument)

   5.  Total No. of CCBs Required:
   (one for each CCV)

   6.  Total No. of Field Blanks Required:
   (one per site or per 10 samples, whichever
   is more frequent)

   7.  Total no. of Lab Blanks Required:
   (one per batch* per method/instrument)

   8.  Total no. of OPR analyses Required:
   (one per batch per method/instrument)

   9.  Total no. of MS/MSD samples Required:
   (one per 10% per matrix per site)

   10.  Total no. Field Duplicates Required:
   (one per 10 samples per site)

   11.  Total no. of MDL results required:
   (one per method and per analyte)	
                                                               No. of Shifts
                     Total No. of CCVs Reported:
                     Total No. of CCBs Reported:
                     Total No. of Field Blanks Reported:
                     Total No. of Lab Blanks Reported:
                     Total No. of OPR Analyses Reported:
                     Total No. of MS/MSD samples Reported:
                     Total No. of Field Duplicates Reported:
                     Total No. of MDL Results Reported:
22
                                                              Draft, July 1996

-------
   12.                                                  Initial Calibration


   a.        Was a multiple point initial calibration performed*?                                      Gyes     Gno

   b.        Were all sample concentrations reported within the calibration range?                              Gyes

            If no, list method and analytes for which initial calibration was not performed or which exceeded the
            calibration range.

   c.        Analyte            No ICAL (Y/N)             Exceeded ICAL Range (Y/N)
   d.        Did the initial calibration meet linearity criteria?                                         Gyes     Gno


   e.        If no, was a calculation curve used to calculate sample concentrations?                    Gyes     Gno

   * A three point (minimum) initial calibration should be performed for each analyte; if the RSD of the mean RRF is less than 15%, or if the RSD of the mean RF i
   less than 25%, then the averaged RRF or RF, respectively, may be used for that analyte.
   13.                                Method Detection Limit (MDL)/Minimum Level (ML)


   a.        Did the laboratory demonstrate their ability to achieve the required MDL?                 Gyes     Gno

   b.        Did the initial calibration range encompass the ML?                                               Gyes

   c.        Were all field samples detected below the ML reported as non-detects?                    Gyes     Gno

   d.        If the answer to item a, b, or c above was "no", describe problem:
   14.                          Initial Calibration Verification (ICV)/Initial Calibration Blanks (ICB):

   a.        Was an ICV run prior to field samples?                                                 Gyes     Gno

   b.        Were ICV results within the specified windows?                                         Gyes     Gno

   c.        Was the ICV followed by an ICB?                                                               Gyes

   d.        Was the ICB free from contamination?                                                  Gyes     Gno

   e.        If any item in a - d above was  answered "no", list problems below:

            Analyte            Failed ICV Recovery         Concentration Detected in ICB         Affected Samples
Draft, July 1996                                                                                                       23

-------
   15.                                        Initial Precision and Recovery (IPR)





   a.        Were IPR data reported for each analyte?                                              Gyes     Gno





   b.        Did all IPR aliquots meet required recovery criteria (x)?                                 Gyes     Gno





   c.        Did the standard deviation (s) of each IPR series meet the required criterion?              Gyes     Gno





   d.        If any item in a - c above was answered "no", document problem below.





            Analyte     Aye. Result Reported (X)            RSD Reported     Affected Samples
   16.                                      Ongoing Precision and Recovery (OPR)





   a.        Were OPR data reported for each analyte, instrument, and batch?                        Gyes     Gno





   b.        Did all OPR samples meet required recovery criteria (x)?                                Gyes     Gno





   c.        If item a or b above was answered "no", document problem below.





            Analyte    OPR Recovery (X) Reported        Shifts Missing OPR         Affected Samples
   17.                    Continuing Calibration Verification (CCV)/Continuing Calibration Blank (CCB)





   a.        Were CCVs run prior to each batch of 10 samples on each instrument?                   Gyes     Gno





   b.        Were all CCV results within the specified windows?                                     Gyes     Gno





   c.        Was each CCV followed by a CCB?                                                   Gyes     Gno





   d.        Was each CCB  free from contamination?                                               Gyes     Gno





   e.        If any item in a - d above was answered "no", list problems below:





            Analyte    Affected Samples       Shift Missing CCV/CCB      Failed CCV/CCB ID
24                                                                                                      Draft, July 1996

-------
18.
a.
b.
c.

19.
a.
b.
c.

20.
a.
b.
c.
d.
e.

Laboratory (Method) Blanks
Was a method blank analyzed for each instrument & sample batch? Gyes Gno
Was each method blank demonstrated to be free from contamination? Gyes Gno
If the answer to item a or b was "no", document problems below.
Analvte Affected Samples

Was a field blank analyzed for each
Was each field blank demonstrated
Blank Concentration Reported Shift Missing MB
Field Blanks
10 samples per site? Gyes Gno
to be free from contamination? Gyes Gno
If the answer to item a or b was "no", document problems below.
Analvte Affected Samples

Blank Concentration Reported Shift Missing FB
MS/MSD Results
Were appropriate number of MS/MSD pairs analyzed? Gyes Gno
Were all MS/MSD recoveries within specified windows? Gyes Gno
Were all RPDs within the specified
window? Gyes Gno
Was appropriate corrective action (e.g., MSA for GFAA, serial dilution
for ICP) employed on affected samples? Gyes Gno
If the answer was "no" to items a -
d above, document affected samples:
Analvte MS % R MSD % R MS/MSD RPD Affected Samples

21.
a.
b.
c.
d.



Additional Information
Were Instrument Tune Data Provided? Gyes Gno
Were equipment blanks demonstrated to be free from contamination? Gyes Gno
Were statements of data quality provided? Gyes Gno
Did field duplicate demonstrate acceptable precision? Gyes Gno
Draft, July 1996                                                                                 25

-------
                                                                                       Glossary
Accuracy:   The  degree of  agreement  between  a measured value  and the  true  or expected  value
of the quantity of concern.

Calibration  Blank:     A  sample  of  reagent  water  analyzed  after  the  calibration  verification
standard to check for contamination attributable to the analytical system.

Calibration  Range  (Calibration  Curve):   A  graphical relationship  between the  known  values  for
a  series  of  calibration  standards  and  instrument   responses,  specifically  the  linear  portion   of
this relationship between calibration standards.

Dissolved Metals:   The  concentration  of metal(s)  that  will pass  through  a  0.45  micron  filter
assembly, prior to acidification  of the sample.

Equipment  Blank:   An  aliquot  of reagent  water that  is  subjected  in  the  laboratory  to  all  aspects
of  sample  collection  and   analysis,   including  contact  with   all  sampling  devices  and  apparatus.
The  purpose of  the   equipment  blank  is  to   determine  if  the  sampling  devices  and  apparatus
for  sample  collection   have  been adequately  cleaned prior to  shipment to  the  field  site.    An
acceptable  equipment  blank   must  be  achieved  before  the   sampling devices  and  apparatus  are
used for sample collection.

Field  Blank:  An aliquot of  reagent water  that  is placed  in  a  sample container in  the  laboratory,
shipped  to  the  sampling  site,  and   treated as  a  sample  in  all  respects,  including  contact  with
the  sampling  devices   and  exposure   to  sampling  site  conditions,  storage,  preservation,  and   all
analytical  procedures,  which  may include  filtration.    The field  blank  is  used  to  determine  if
field   sample  handling  processes,   sample   transport,   and   sampling   site   environment  have
caused sample contamination.

Field  Duplicates:   Two  identical aliquots of  a  sample  collected  in  separate  sample  containers
at the same  time  and  place  under   identical  circumstances  and  sample   collection  techniques,
and  handled  in  exactly  the  same manner as  other  samples.    Field  duplicates  are  used  as a
measure  of   the   precision   associated  with  sample  handling, preservation,  and  storage  as   well
as laboratory handling, preparation, and analytical procedures.

Initial  Precision  and  Recovery (IPR):   A  series  of four consecutively analyzed  aliquots  of reagent
water  containing  the analyte(s) of interest at 2 -  3 times the  ML.    IPRs   are  performed  prior
to the first  time  a  method  is  used  and any  time  the  method  or  instrumentation  is  modified.
The    IPR  is  used   to  demonstrate   the  analyst/laboratory   ability   to   generate   acceptable
precision  and  accuracy through   the  calculated  mean  (x)  and  standard  deviation   (s)  for  each
analyte.

Laboratory  Blank:   An  aliquot  of  reagent water  that   is treated  exactly  as a sample  including
exposure   to   all  glassware,   equipment,  solvents,   reagents,   internal  standards,  and  surrogates
that   are   used  with   samples.    the   laboratory  blank   is   used   to   determine  if  analytes   or
interferences are present in the  laboratory environment, reagents, or the apparatus.
26                                                                                   Draft, July 1996

-------
Magnetic Media:    A  storage medium  on  which  all  instrumentally  acquired  raw  data  may be
retained.

Matrix Spike  (MS)  and Matrix Spike Duplicate (MSD):   Aliquots of an environmental sample  to
which   known  quantities  of analytes   are  added  in   the  laboratory.    The   MS  and  MSD  are
analyzed  under  the  same   conditions  as  other  samples  and  are  used  to  quantify  the  bias  and
precision  associated  with   the  sample   matrix.    The  background  concentration  of  the   analytes
in the sample are determined and subtracted from the MS and MSD results.

Method Blank: See "laboratory blank".

Method  Detection  Limit  (MDL):   The  minimum  concentration of  an analyte that,  in  a  given
matrix  and  with  a  specified  method,  has  a  99%  probability  of  being identified,  qualitatively
or quantitatively measured, and reported to be greater than zero.

Minimum  Level   (ML):     The  lowest  level   at  which  the  entire  analytical  system  gives   a
recognizable signal and acceptable calibration point.

Ongoing Precision  and  Recovery  (OPR):  An  aliquot of reagent water containing the analyte(s)
of  interest.    The  OPR  is  used  to  demonstrate  continuing  ability  of  the   analyst/laboratory  to
generate acceptable results based on target and standard recoveries.

Quality Assurance  (QA):    An  integrated  system of  activities  involving planning,  quality  control,
quality  assessment,   reporting,  and  quality  improvement  to  ensure   that  a  product   or  service
meets defined standards of quality with  a stated level of confidence.

Quality Control  (QC):   The  overall  system  of  technical activities designed  measure and  control
the  quality of a product  or  service  so  that it  meets  the needs  of users.   The  aim  is  to  provide
quality that is satisfactory, adequate, dependable, and economical.

Precision:    The   degree   of  mutual   agreement   characteristic   of independent   measurements  as
the  result of repeated applications  of the process under specified conditions.

Reagent  Water:   Water  demonstrated  to  be  free  from  the metal(s)  of  interest  at the  method
detection  limit  (MDL)  of  the  analytical  method  to  be  used  for determination  of the   metal(s)
of interest.

Reference Standards:  A  material or  substance, one or  more  properties of  which  are  sufficiently
well  established  to  be  used  for  the  calibration of analytical  apparatus,  the   assessment  of  a
measurement method, or assigning of values to materials.

Trace  Metals:    Concentrations  of  metals  found   at   or   near their  established  water  quality
criteria levels.
Draft, July 1996                                                                                   27

-------
                                                                                Appendix A

                                                                EPA Water Quality Criteria for
                                                   Priority Pollutant Metals and Metals Species
The  table  provided  on  the  following  page  provides  the  freshwater,  marine,  and  human  health
water  quality  criteria  published  by  EPA  for  priority  pollutant  metals  and   metals   species.
Human health criteria  reflect values published  by EPA in  the  National Toxics  Rule  at  57  PR
60848.   Aquatic  criteria  reflect  values  published by EPA in the  National Toxics  Rule  and  in
the  Stay  of  Federal  Water  Quality  Criteria for  Metals  (60  PR  22228).    This  table  includes
criteria  for  both  total  recoverable metals  and  dissolved metals.    In  addition,  the  table  includes
freshwater  criteria  that are  based on a hardness of 100  mg/L.   In  order to  provide a  worst-
case  scenario,  the  table  also  includes criteria that  are  based on  a hardness  of 25  mg/L  CaCO3.
Calculations for deriving these values were published by EPA at 60 PR 22228.
28                                                                                 Draft, July 1996

-------
                EPA Ambient Water Quality Criteria for Total Recoverable and Total Dissolved Priority Pollutant Metals  and Metal Species
                                                        Calculated at a Hardness of 100 mg/L and 25 mg/L CaCO3
Metal
Sb
As
Cd'6'
Cr (III)'6'
Cr (VI)
Cu'''
pbW
Hg
Ni<6'
Se
Ag<"
TI
Zn<6'
Ambient Water Quality Criteria'1' (re/L)
Freshwater Criteria
Acute'2'
Tot. Rec.
100 mg/L
CaCOj

360
3.9
1700
16
18
82
2.4
1400
20
4.1

120
Acute'3'
Tot. Diss.
100 mg/L
CaCO3

360
3.7
550
15
17
65
2.1
1400
___P)
3.4

110
Acute<4>
Tot. Rec.
25 mg/L
CaCOj

360
0.82
560
16
4.8
14
2.4
440
20
0.37

36
Acute'3"4'
Tot. Diss.
25 mg/L
CaCO3

360
0.82
180
15
4.6
14
2.1
440
___P)
0.32

35
Chrome'2'
Tot. Rec.
100 mg/L
CaCOj

190
1.1
210
11
12
3.2
0.012
160
5.0


110
Chrome'3'
Tot. Diss.
100 mg/L
CaCO3

190
1.00
180
10
11
2.5
	 (?)
160
___P)


100
Chronic'4'
Tot. Rec.
25 mg/L
CaCOj

190
0.38
67
11
3.6
0.54
0.012
49
5.0


33
Chronic'3'-'4'
Tot. Diss.
25 mg/L
CaCOj

190
0.37
57
10
3.5
0.54
	 (?)
49
___P)


32
Marine Criteria
Acute'2'
Tot. Rec.


69
43

1100
2.9
220
2.1
75
290
2.3

95
Acute'3'
Tot. Diss.


69
42

1100
2.4
210
1.8
74
290
2.0

90
Chronic'2'
Tot. Rec.


36
9.3

50
2.9
8.5
0.025
8.3
71


86
Chronic'3'
Tot. Diss.


36
9.3

50
2.4
8.1
(7)
8.2
71


81
Human Health Criteria
H2O /organism'2'
Tot. Rec.
ws>
O.OttP>





0.14
610'5'


1.W

organism'3'
Tot. Rec.
4300'5'
0.14'5'





0.15
4600'5'


6.3'5'

WQC promulgated at 40 CFR Part 131 (57 FR 60848 and 60 FR 22228).  Cntna for metals listed at 40 CFR Part 131 are expressed as total recoverable at a hardness of 100 mg/L CaCO3 and a water effect ratio (WER) of
1.0. The lowest WQC for each analyte is shaded.

As listed in the NTR at 40 CFR Part 131 for total recoverable metals. Hardness dependent freshwater acute and chronic criteria expressed at a hardness of 100 mg/L CaCO3 and a WER of 1.0.

For As, Cd, Cr(III), Cr(VI), Cu, Ni, Pb, and Zn, acute and chronic criteria for dissolved metals and metal species were calculated in accordance with the equations provided at 60 FR 22228. Hardness-dependent dissolved
criteria conversion factors for Cd and Pb also cacluated at a hardness of 25 mg/L per 60 FR 22228.

Hardness dependent freshwater acute and chronic criteria recalculated at a hardness of 25 mg/L CaCO3 and a WER of 1.0 as specified at 40 CFR Part 131.36 (b)(2). For dissolved metals, hardness calculations were performed
prior to adjusting for dissolved levels.

Criterion reflects recalculated value using IRIS.

Freshwater criteria are hardness dependent for this metal.

Metal is bio accumulative and, therefore, it is not appropriate to calculate WQC for dissolved levels. (Guidance Document on Dissolved Criteria:  Expression oJA.quatic Life Criteria, October 1993.  Attachment 2 to memorandum
from Martha Pro thro to Water Management Division Directors, October 1, 1993.)

-------