EPA Document # EPA 815-B-01-001
METHOD 317.0  DETERMINATION OF INORGANIC OXYHALIDE DISINFECTION
               BY-PRODUCTS IN DRINKING WATER USING ION
               CHROMATOGRAPHY WITH THE ADDITION OF A
               POSTCOLUMN REAGENT FOR TRACE BROMATE ANALYSIS

                                Revision 2.0

                                July 2001
Herbert P. Wagner and Barry V. Pepich, IT Corporation and Daniel P. Hautman and
David J. Munch, US EPA, Office of Ground Water and Drinking Water
                      TECHNICAL SUPPORT CENTER
            OFFICE OF GROUND WATER AND DRINKING WATER
               U. S. ENVIRONMENTAL PROTECTION AGENCY
                         CINCINNATI, OHIO 45268
                                 317.0-1

-------
                                   METHOD 317.0

 DETERMINATION OF INORGANIC OXYHALIDE DISINFECTION BY-PRODUCTS
 IN DRINKING WATER USING ION CHROMATOGRAPHY WITH THE ADDITION
        OF A POSTCOLUMN REAGENT FOR TRACE BROMATE ANALYSIS
1. SCOPE AND APPLICATION

   1.1  This method covers the determination of inorganic oxyhalide disinfection by-product
        anions in reagent water, surface water, ground water, and finished drinking water. In
        addition, bromide can be accurately determined in source or raw water and it has been
        included due to its critical role as a disinfection by-product precursor.  Bromide
        concentration in finished water can differ significantly between preserved and
        unpreserved samples and should not be attempted due to numerous variables which can
        influence the concentration. Since this method, prior to the addition of the postcolumn
        reagent (PCR), employs the same hardware as EPA Method 300.11, the analysis of the
        common anions (using EPA Method 300.1, Part A1) can be performed using this
        instrument setup with the postcolumn hardware attached but "off-line" and with the
        appropriate smaller sample loop.

	Inorganic Disinfection By-products by Conductivity Detection
        Bromate (report values > 15.0 ug/L)             Chlorite
        Bromide (source or raw water only)             Chlorate

	Inorganic Disinfection By-product by Postcolumn UV/VIS Absorbance Detection
        Bromate (report values > Minimum Reporting Limit (MRL) to  15.0 ug/L)

   1.2  The single laboratory reagent water Method Detection Limits (MDL, defined in Section
        3.14) for the above analytes are listed in Table 1.  The MDL for a specific matrix may
        differ from those listed, depending upon the nature of the sample and the specific
        instrumentation employed.

        1.2.1   In order to achieve comparable detection limits on the conductivity detector, an
               ion chromatographic system must utilize suppressed conductivity detection, be
               properly maintained and must be  capable of yielding a baseline with no more
               than 5 nanosiemen (nS) noise/drift per minute of monitored response over the
               background conductivity.

        1.2.2  In order to achieve acceptable detection limits on the postcolumn absorbance
               detector, the postcolumn reagent must be delivered pneumatically and some
               form of software signal filtering or smoothing of the absorbance signal from the
               absorbance detector must be incorporated.2
                                       317.0-2

-------
    1.3  This method is recommended for use only by or under the supervision of analysts
        experienced in the use of ion chromatography and in the interpretation of the resulting
        ion chromatograms.

    1.4  When this method is used to analyze unfamiliar samples for any of the above anions,
        anion identification should be supported by the use of a fortified sample matrix
        covering the anions of interest.  The fortification procedure is described in Section
        9.4.1.

    1.5  Users of the method data should state the  data quality objectives prior to analysis.
        Users of the method must demonstrate the ability to generate acceptable results with
        this method, using the procedures described in Section 9.0.

 2.  SUMMARY OF METHOD

    2.1  A volume of sample, approximately 225  |_iL (see Note), is introduced into an ion
        chromatograph (1C). The anions of interest are separated and measured, using a system
        comprised of a guard column, analytical column,  suppressor device, conductivity
        detector, a postcolumn reagent delivery system  (pneumatically controlled), a heated
        postcolumn reaction coil, and a ultraviolet/visible (UVYVIS) absorbance detector.2'3

        NOTE:  A 225 uL sample loop can be made using approximately 111 cm (44 inches)
                 of 0.02  inch i.d. PEEK tubing.  Larger injection loops may be employed.4
                 The volume should be verified to be within 5% by weighing the sample loop
                 empty, filling the loop with deionized water and re-weighing the loop
                 assuming the density of water is 1 mg/uL.

3.   DEFINITIONS

    3.1  ANALYSIS BATCH - A sequence of samples, which are analyzed within a 30 hour
        period and include no more than 20 field samples. An analysis batch must also include
        all required QC samples, which do not contribute to the maximum field sample total of
        20. The required QC samples include:
        • Laboratory Reagent Blank (LRB)
        • Initial Calibration Check Standard (ICCS)
        • Laboratory Fortified Blank (LFB)
        • Continuing  Calibration Check Standard (CCCS), when the batch contains more than
          10 field samples
        • End Calibration Check Standard (ECCS)
        • Laboratory Fortified Matrix (LFM)
        • Either a Field Duplicate (FD), a Laboratory Duplicate (LD) or a duplicate of the LFM

    3.2  CALIBRATION STANDARD (CAL) - A solution prepared from the primary dilution
        standard solution(s) or stock standard solutions and the surrogate analyte. The CAL

                                        317.0-3

-------
     solutions are used to calibrate the instrument response with respect to analyte
     concentration.

3.3  INITIAL CALIBRATION STANDARDS - A series of CAL solutions (either
     individual or combined target analytes) used to initially establish instrument calibration
     and develop calibration curves for individual target anions (Section 10.2).

3.4  INITIAL CALIBRATION CHECK STANDARD (ICCS) - A CAL solution, (either
     individual or combined target analytes) which is analyzed initially, prior to any field
     sample analyses, which verifies previously established calibration curves. The
     concentration for the initial calibration check standard MUST be at or below the MRL
     (Section 3.15) level which is also the level of the lowest calibration standard (Section
     10.3.1).

3.5  CONTINUING CALIBRATION CHECK STANDARDS (CCCS) - A CAL solution
     (either individual or combined target analytes) which is analyzed after every tenth field
     sample analyses which verifies the previously established calibration curves and
     confirms accurate analyte quantitation for the previous ten field samples analyzed. The
     concentration for the continuing calibration check standards should be either at a
     middle calibration level or at the highest calibration level (Section 10.3.2).

3.6  END CALIBRATION CHECK STANDARD  (ECCS) - A CAL solution (either
     individual or combined target analytes) which is analyzed after the last field sample
     analysis which verifies the previously established calibration curves and confirms
     accurate analyte quantitation for all field samples analyzed since the last continuing
     calibration check. The end calibration check standard should be either the middle or
     high level continuing calibration check standard (Section 10.3.2).

3.7  FIELD DUPLICATES (FD) - Two separate field samples collected at the same time
     and place under identical circumstances and handled exactly the same throughout field
     and laboratory procedures. Analyses of field duplicates indicate the precision
     associated with sample collection, preservation and storage, as well as with laboratory
     procedures.

3.8  INSTRUMENT PERFORMANCE CHECK SOLUTION (PC) - A solution of one or
     more method analytes, surrogates, or other test substances used to evaluate the
     performance of the instrument system with respect to a defined set of criteria.

3.9  LABORATORY DUPLICATE (LD) - Two sample aliquots, taken in the laboratory
     from a single field sample bottle, and analyzed separately with identical procedures.
     Analysis of the initial sample (LJ and the duplicate sample [(Dc) Section 9.4.3.1]
     indicate precision associated specifically with the laboratory procedures by removing
     variation contributed from sample collection, preservation and storage procedures.
                                    317.0-4

-------
3.10 LABORATORY FORTIFIED BLANK (LFB) - An aliquot of reagent water or other
     blank matrix to which known quantities of the method analytes are added in the
     laboratory.  The LFB is analyzed exactly like a sample, and its purpose is to determine
     whether the methodology is in control, and whether the laboratory is capable of making
     accurate and precise measurements.

3.11 LABORATORY FORTIFIED SAMPLE MATRIX (LFM) - An aliquot of an
     environmental field sample to which known quantities of the method analytes  are added
     in the laboratory. The LFM is analyzed exactly like a sample, and its purpose  is to
     determine whether the field sample matrix contributes bias to the analytical result. The
     background concentrations of the analytes in the field sample matrix must be
     determined in a separate, unfortified aliquot and the measured values in the LFM
     corrected for background concentrations.

3.12 LABORATORY REAGENT BLANK (LRB) - An aliquot of reagent water or other
     blank matrix that is handled exactly as a sample including exposure to all glassware,
     equipment, solvents, reagents, and surrogates that are used with other samples. The
     LRB is used to determine if method analytes or other interferences are present in the
     laboratory environment, the reagents,  or the apparatus.

3.13 MATERIAL SAFETY DATA SHEET (MSDS) - Written information provided by
     vendors concerning a chemical's toxicity, health hazards, physical properties, fire, and
     reactivity data including storage, spill, and handling precautions.

3.14 METHOD DETECTION LIMIT (MDL) - The minimum concentration of an analyte
     that can be identified, measured and reported with 99% confidence that the analyte
     concentration is greater than zero.5

3.15 MINIMUM REPORTING LEVEL (MRL) - The minimum concentration that can be
     reported as a quantitated value for a target analyte in a sample following analysis. This
     defined concentration can be no lower than the concentration of the lowest calibration
     standard and can only be used if acceptable quality control criteria for the ICCS are met.

3.16 PROFICIENCY TESTING (PT) or PERFORMANCE EVALUATION (PE) SAMPLE
     - A certified solution of method analytes whose concentration is unknown to the
     analyst. Frequently, an aliquot of this solution is added to a known volume of reagent
     water and analyzed with procedures used for samples. Often,  results of these analyses
     are used as part of a laboratory certification program to objectively determine  the
     capabilities of a laboratory to achieve high quality results.

3.17 QUALITY CONTROL SAMPLE (QCS) - A solution of method analytes of known
     concentrations that is obtained from a source external to the laboratory and different
     from the source  of calibration standards. It is used to check laboratory performance
     with externally prepared test materials.

                                   317.0-5

-------
   3.18  SURROGATE ANALYTE - An analyte added to all samples, standards, blanks, etc.,
         which is unlikely to be found at a significant concentration, and which is added directly
         in a known amount before any sample processing procedures are conducted (except in
         the procedure for the removal of chlorite as described is Section 11.1.4). It is measured
         with the same procedures used to measure other sample components. The purpose of
         the surrogate analyte is to monitor method performance with each sample.

   3.19  STOCK STANDARD SOLUTION (SSS) - A concentrated solution containing  one or
         more method analytes prepared in the laboratory using assayed reference materials or
         purchased from a reputable commercial source.

4. INTERFERENCES

   4.1   Interferences can be divided into three different categories: direct chromatographic
         coelution, where an analyte response is observed at very nearly the same retention time
         as the target anion; concentration dependant coelution, which is observed when the
         response of higher than typical concentrations of the neighboring peak overlap into the
         retention window of the target anion; and, ionic character displacement, where retention
         times may significantly shift due to the influence of high ionic strength matrices  (high
         mineral content or hardness) overloading  the exchange sites on the column and
         significantly shortening target analyte's retention times.

         4.1.1   A direct chromatographic coelution may be solved by changing columns, eluent
               strength, modifying the eluent with organic solvents (if compatible  with 1C
               columns), changing the detection systems, or selective  removal of the
               interference with pretreatment. Sample dilution will have little to no effect.
               The analyst must verify that these changes do not induce any negative affects on
               method performance by repeating  and passing all the QC criteria as described in
               Section 9.

         4.1.2  Sample dilution may resolve some of the difficulties if the interference is the
               result of either concentration dependant coelution or ionic character
               displacement,  but it must be clarified that sample dilution will alter your
               Minimum Reporting Limit (MRL) by a proportion equivalent to that of the
               dilution. Therefore, careful  consideration of project objectives should be given
               prior to performing such a dilution.  An alternative to sample dilution, may be
               dilution of the eluent as outlined in Section 11.2.6.

         4.1.3   Pretreatment cartridges can be effective as a means to eliminate certain matrix
               interferences.  With any proposed pretreatment, the analyst must verify that
               target analyte(s) are not affected by monitoring recovery after pretreatment.
               With advances in analytical separator column technology which employ higher
               capacity anion exchange resins, the need for these cartridges has been greatly
               reduced.

                                        317.0-6

-------
   4.2  Method interferences may be caused by contaminants in the reagent water, reagents,
        glassware, and other sample processing apparatus that lead to discrete artifacts or
        elevated baselines in an ion chromatogram. These interferences can lead to false
        positive results for target analytes as well as reduced detection limits as a consequence
        of elevated baseline noise.

   4.3  Samples that contain particles larger than 0.45 microns and reagent solutions that
        contain particles larger than 020 microns require filtration to prevent damage to
        instrument columns and flow systems.

   4.4  Close attention should be given to the potential for carry over peaks from one analysis
        which will effect the proper detection of analytes of interest in a second or subsequent
        analysis. Normally, in this analysis, the elution of sulfate (retention time of 17.5 min.)
        indicates the end of a chromatographic run, but, in the ozonated and chlorine dioxide
        matrices, which were included as part of the single operator accuracy and bias study,  a
        small response (200 nS baseline rise) was observed for a very late eluting unknown
        peak following the response for sulfate. Consequently, a run time of 25 minutes is
        recommended to allow for the proper elution of any potentially interferant late peaks. It
        is the responsibility of the user to confirm that no late eluting peaks have carried over
        into a subsequent analysis thereby compromising the integrity of the analytical results.

   4.5  Any residual chlorine dioxide present in the sample will result in the formation of
        additional  chlorite prior to analysis. If residual chlorine dioxide is suspected in the
        sample, the sample must be purged with an inert gas (helium, argon or nitrogen) for
        approximately five minutes. This sparging must be conducted prior to ethylenediamine
        preservation and at the time of sample collection.

   4.6  The presence of chlorite can interfere with the quantitation of low concentrations of
        bromate on the postcolumn UV/VIS absorbance detector.  In order to accurately
        quantify bromate concentrations in the range 0.5 - 15.0  |-ig/L in this postcolumn system,
        the excess chlorite must be removed prior to analysis as outlined in Section 11.1.4.

5.  SAFETY

   5.1  The toxicity or carcinogenicity of each reagent used in this method have not been fully
        established although the postcolumn reagent o-dianisidine, is listed as a potential
        human carcinogen. Each chemical should be regarded as a potential health hazard and
        exposure should be as low as reasonably achievable. Cautions are included for known
        extremely  hazardous materials or procedures.

   5.2  Each laboratory is responsible for maintaining a current awareness file of Occupational
        Safety and Health Administration (OSHA) regulations regarding the safe handling of
        the chemicals specified in this method. A reference file of Material Safety Data Sheets
        (MSDS) should be made available to all personnel involved in the chemical analysis.

                                        317.0-7

-------
        The preparation of a formal safety plan is also advisable.  Additional references on
        laboratory safety are available.6"9

   5.3  The following chemicals have the potential to be highly toxic or hazardous, consult
        MSDS.

        5.3.1   Sulfuric acid - used to prepared a 25 mN sulfuric acid regenerant solution for
               chemical suppression using a Dionex Anion Micro Membrane Suppressor
               (AMMS) and for pretreatment for chlorite removal (Section 11.1.4)

        5.3.2   Nitric acid - used to prepare the postcolumn reagent.

        5.3.3   o-dianisidine [3, 3'- dimethoxybenzidine dihydrochloride (ODA)] - used as the
               postcolumn reagent.

6.  EQUIPMENT AND SUPPLIES

   6.1  Ion chromatograph - Analytical system complete with ion chromatographic pump and
        all required accessories including syringes, analytical columns, compressed gasses,
        suppressor, conductivity detector, mixing "tee", postcolumn reagent delivery system,
        reaction coil, reaction coil heater, UV/VIS absorbance detector (Figure 1) and a PC
        based data acquisition and control system.

        NOTE: Because of its acidic nature and high salt content,  the PCR MUST be flushed
        from the reaction coil upon completion of the final analysis and prevented from
        draining through the reaction coil by gravity once the system is shut down. This can be
        accomplished either manually or by incorporating a column switching valve in
        combination with a flush and close method in the schedule.

        6.1.1   Anion guard column - Dionex AG9-HC 4 mm (P/N 51791), or equivalent.
               This column functions as a protector of the separator column.  If omitted from
               the system the retention times will be shorter.

        6.1.2   Anion separator column - Dionex AS9-HC column, 4 mm (P/N 51786), or
               equivalent (see Note). The AS9-HC, 4 mm column using the conditions outlined
               in Table 1 produced the separations shown in Figures 2 and 3.

               NOTE: The use of 2 mm columns is not recommended. A 50 uL sample loop
               would be required with the 2 mm columns. This reduced injection volume
               would decrease the "on-column" bromate and negatively affect PCR reactivity
               and the subsequent absorbance response. As well,  the 2 mm columns require a
               flow rate approximately 4 times less than the 4 mm columns. At the lower flow
               rates, band broading may become an issue and it would be difficult, if not
               impossible, to accurately maintain the appropriate reduced flow rate for the
               PCR.

                                       317.0-8

-------
6.1.3   Anion suppressor device - The data presented in this method were generated
       using a Dionex Anion Self Regenerating Suppressor (4 mm ASRS, P/N 46081).
       An equivalent suppressor device may be utilized provided comparable
       conductivity detection limits are achieved and adequate baseline stability is
       attained as measured by a combined baseline drift/noise of no more than 5 nS
       per minute over the background conductivity. The suppressor must be able to
       withstand approximately 80 - 120 psi back pressure which results from
       connecting the postcolumn hardware to the eluent out side of the suppressor.

       6.1.3.1   The ASRS was set to perform electrolytic suppression at a current
               setting of 100 mA using the external water mode. Insufficient baseline
               stability was observed on the conductivity detector using the ASRS in
               recycle mode.

       6.1.3.2  This method was developed as a multiple component procedure
               employing both suppressed conductivity and postcolumn UV/VIS
               absorbance detectors in series.  If a laboratory is exclusively interested
               in monitoring trace bromate using the PCR and the UV/VIS
               absorbance detector, the suppressor may not be required. The
               performance data presented within this method for the PCR and
               UV/VIS absorbance detector, is based upon a suppressed mobile phase
               system. A laboratory must generate comparable data as a result of a
               complete IDC (Section  9.2) in order to demonstrate comparability of a
               non suppressed system.

6.1.4   Detector - Conductivity cell (Dionex CD20, or equivalent) capable of providing
       data as required in Section 9.2.

6.1.5   Detector - Absorbance detector (Dionex AD20 or equivalent with 10 mm cell
       pathlength, equipped with a tungsten source bulb, or equivalent and capable of
       measuring absorbance at 450 nm) capable of providing data as required in
       Section 9.2.

6.1.6   Postcolumn reagent delivery system (Dionex PC-10, or equivalent),
       pneumatically delivers the postcolumn reagent to mixing tee. The pressure
       settings will need to be established on an individual basis for each specific
       instrument configuration and at a level which yields the prescribed PCR flow
       rates.

6.1.7   Reaction Coil, 500 uL internal volume, knitted, potted or configured to fit
       securely in the postcolumn reaction coil heater. (Dionex P/N 39349, or
       equivalent).
                               317.0-9

-------
        6.1.8  Postcolumn Reaction Coil Heater, capable of maintaining a temperature of up to
               80°C.  (Dionex PCH-2, or equivalent).

   6.2  Data System - The Dionex Peaknet Data Chromatography Software was used to
        generate all the data in the attached tables. Other computer based data systems may
        achieve approximately the same MDLs but the user must demonstrate this by the
        procedure outlined in Section 9.2.

   6.3  Analytical balance - Used to accurately weigh target analyte salts for stock standard
        preparation (±0.1 mg sensitivity).

   6.4  Top loading balance  - Used to accurately weigh reagents to prepare eluents (±10 mg
        sensitivity).

   6.5  Weigh boats - Plastic, disposable - for weighing eluent reagents.

   6.6  Syringes - Plastic, disposable, 10 mL - used during sample preparation.

   6.7  Pipets -  Pasteur, plastic or glass, disposable, graduated, 5 mL and 10 mL.

   6.8  Bottles - High density polyethylene (HOPE), opaque or glass, amber, 30 mL, 125 mL,
        250 mL, used  for sample collection and storage of calibration solutions.  Opaque or
        amber bottles  are required due to the photoreactivity of the chlorite anion.

   6.9  Micro beakers - Plastic, disposable - used during sample preparation.

   6.10 Particulate filters - Gelman ion chromatography Acrodisc 0.45 micron (PN 4485)
        syringe filters  or equivalent. These cartridges are used to remove particulates and
        [Fe(OH)3(s)] which is formed during the oxidation-reduction react ion between Fe (II)
        and C1O2-.

   6.11 Hydrogen cartridges  - Dionex OnGuard-H cartridges (PN 039596) or equivalent.
        These cartridges are conditioned according to the manufacturer's directions and are
        used to protect the analytical column and the suppressor membrane by removing excess
        ferrous iron [Fe (II)]. The ferrous iron is added to field samples to reduce chlorite
        levels prior to analysis of chlorine dioxide disinfected water samples.

7.  REAGENTS AND STANDARDS

   7.1  Reagent water- Distilled or deionized water 18  M Q or better, free of the anions of
        interest.  Water should contain particles no larger than 0.20 microns.

   7.2  Eluent solution - Sodium carbonate (CASRN 497-19-8) 9.0 mM. Dissolve 1.91  g
        sodium carbonate (NajC^) in reagent water and dilute to 2 L.

                                       317.0-10

-------
     7.2.1   This eluent solution must be purged for 10 minutes with helium prior to use to
            remove dissolved gases which may form micro bubbles in the 1C compromising
            system performance and adversely effecting the integrity of the data.
            Alternatively, an in-line degas apparatus maybe employed.

7.3  Stock standard solutions, 1000 mg/L (1 mg/mL) - Stock standard solutions maybe
     purchased as certified solutions or prepared from ACS reagent grade, potassium or
     sodium salts as listed below, for most analytes. Chlorite requires careful consideration
     as outlined below in Section 7.3.4.1.

     7.3.1   Bromide (Br) 1000 mg/L -  Dissolve 0.1288 g sodium bromide (NaBr, CASRN
            7647-15-6) in reagent water and dilute to  100 mL in a volumetric flask.

     7.3.2   Bromate (BrO3~) 1000 mg/L - Dissolve 0.1180 g of sodium bromate (NaBrO3,
            CASRN 7789-38-0) in reagent water and dilute to 100 mL in a volumetric flask.

     7.3.3   Chlorate (C1CV) 1000 mg/L - Dissolve 0.1275 g of sodium chlorate (NaClO3,
            CASRN 7775-09-9) in reagent water and dilute to 100 mL in a volumetric flask.

     7.3.4   Chlorite (C1O2~) 1000 mg/L - If the amperometric titration of the technical
            grade sodium chlorite (NaClO2), specified in 7.3.4.1, had indicated the purity of
            the salt to be 80.0 % NaClO2, the analyst would dissolve 0.1676 g of sodium
            chlorite (NaClO2, CASRN 7758-19-2) in reagent water and dilute to 100 mLin
            a volumetric flask.

            7.3.4.1   High purity sodium chlorite (NaCIO 2) is not currently commercially
                    available due to its potential explosive instability.  Recrystallization of
                    the technical grade (approx. 80%) can be performed but it is labor
                    intensive and time consuming. The simplest approach is to determine
                    the exact purity of the NaCIO 2 using the iodometric titration
                    procedure.10 Following titration, an individual  component standard of
                    chlorite must be analyzed to determine if there  is any significant
                    contamination (greater than 1% of the chlorite weight) from chlorate,
                    bromate or bromide (as other method target anions) in the technical
                    grade chlorite standard.

                    NOTE: Stability of standards - Stock standards (Section 7.3) for most
                    anions are stable for at least 6 months when refrigerated at <6°C.  The
                    chlorite standard is only stable for two weeks when stored refrigerated
                    at <6°C and protected from light. Dilute working standards should be
                    prepared monthly, except those that contain chlorite, which must be
                    prepared every two weeks or sooner if signs of degradation are
                    indicated by repeated QC failure.
                                    317.0-11

-------
7.4  Ethylenediamine (EDA) preservation solution, 100 mg/mL - Dilute 2.8 mL of
     ethylenediamine (99%) (CASRN 107-15-3) to 25 mL with reagent water.  Prepare fresh
     monthly.

7.5  Surrogate Solution, 0.50 mg/mL dichloroacetate (DCA) - Prepare by dissolving 0.065 g
     dichloroacetic acid, potassium salt (C12CHCO2K, CASRN 19559-59-2) in reagent water
     and diluting to 100 mL in a volumetric flask.

       7.5.1   Dichloroacetate is potentially present in treated drinking waters as the acetate
              of the organic disinfection by product, dichloroacetic acid (DCAA). Typical
              concentrations of DCAA rarely exceed 50 |-ig/L, which, for this worst case
              example, would represent only a five percent increase in the observed
              response over the fortified concentration of 1.00 mg/L.  Consequently, the
              criteria for acceptable recovery (90% to 115%) for the surrogate is weighted to
              115% to allow for this potential background.

     7.5.2   Prepare this solution fresh every 3 months or sooner if signs of degradation are
            indicated by the repeated failure of the surrogate QC criteria.

     7.5.3   If the analyst is exclusively interested  in monitoring trace bromate using the
            PCR and the UV/VIS absorbance detector, the surrogate may be omitted since it
            only yields a signal on the conductivity detector. If the surrogate is removed,
            the laboratory must adhere to the alternate QC requirements found in Section
            9.3.3.3 in order to monitor and demonstrate proper instrument performance.

7.6  Postcolumn reagent - The postcolumn reagent is prepared by adding 40 mL of 70%
     redistilled nitric acid (purity as 99.999+%, Aldrich, Cat. No. 22,571-1, Milwaukee, WI,
     or equivalent) to approximately 300 mL reagent water (see Note  1) in a well rinsed 500
     mL volumetric flask (see Note 2) and adding  2.5 grams of ACS reagent grade KBr
     (Sigma, Cat. No. P-5912, St. Louis, MO, or equivalent). Two-hundred-and-fifty
     milligrams of purified grade o-dianisidine, dihydrochloride salt [(ODA), (Sigma, Cat.
     No. D-3252, or equivalent)] are dissolved, with stirring, in 100 mL methanol
     (Spectrophotometric grade, Sigma, Cat. No. M-3641, St. Louis MO, or equivalent).
     After dissolution,  the o-dianisidine solution is added to the nitric  acid/KBr solution and
     diluted to volume with reagent water. The reagent is stable  for 24 hours and should be
     prepared fresh daily prior to analysis.

     7.6.1   The purity of all reagents  employed in the preparation of the postcolumn reagent
            is critical.  Some commercial manufacturers/suppliers of laboratory chemicals
            sell inferior grades of o-dianisidine dihydrochloride.  ONLY the purified grade
            of this reagent is acceptable (see Notes 3 and 4). The purified ODA
            dihydrochloride salt is a white, fine powder.
                                    317.0-12

-------
               NOTE 1: For selected lots of ODA, the method sensitivity monitored by the
               UV-vis detector may be increased by as much as 2-fold if the reagent water used
               to prepare the ODA PCR is purged with helium for 30 minutes prior to
               preparing the ODA solution.

               NOTE 2: All glassware used to prepare the postcolumn reagent must be
               thoroughly rinsed with reagent water prior to use. A champagne or light amber
               coloration of the PCR reagent may be evident when freshly prepared. Over
               several hours, this slight coloration will fade. Consequently, the PCR must be
               prepared in advance and allowed to sit until it is clear, for a minimum of 4 hours
               (preferably  overnight) prior to use. Occasionally, no matter how well all the
               glassware used to prepare the postcolumn reagent is rinsed, a darkly colored
               solution (oxidized ODA) may result. These solutions MUST be discarded. For
               this reason, it is recommended that the PCR be made in a series of 500 mL lots
               with dedicated glassware. The clear solution should be filtered using a 0.45
               micron membrane to remove particulates before use.

               NOTE 3: Differences in purity as indicated by variations in the physical
               appearance of different lots of ODA, even from the same manufacturer can
               effect method sensitivity.  Although considerably more expensive, a pelletized
               form of ODA from one supplier (Sigma, Cat. No. D-9154, St. Louis, MO) has
               shown to increase method sensitivity by as much as 2-fold over impure lots of
               ODA. Care must be exercised when switching ODA lots to ensure the method
               sensitivity is not compromised.

               NOTE 4: The PCR reaction temperature was optimized at 60 °C with the
               granular ODA that was available during the original method development.
               Investigation of recent changes in physical appearance/decreased sensitivity and
               stability of the ODA PCR indicated that, the method sensitivity with the ODA
               currently available, can be dramatically increased (up to 180%) or increased by
               up to a factor of 1.8) by increasing the reaction temperature to 80 °C. Use of
               temperatures ranging from 60 to 80 °C may be used in this method.

   7.7  Ferrous iron [1000 mg/L Fe (II)] solution - Dissolve 0.124 g ferrous sulfate
        heptahydrate (FeSO4.7H2O, Sigma, F-7002) in approximately 15 mL reagent water
        containing 6 uL concentrated nitric acid and dilute to 25 mL with reagent water in a
        volumetric flask (final pH ~2). The Fe (II) solution must be prepared fresh every two
        days.

   7.8  Sulfuric acid (0.5 N) - Dilute  1.4 mL of concentrated sulfuric acid (Fisher Scientific
        Certified ACS Plus, A 3 00-500) to 100 mL.

8.  SAMPLE COLLECTION,  PRESERVATION AND STORAGE


                                       317.0-13

-------
 8.1   Samples should be collected in plastic or glass bottles. All bottles must be thoroughly
      cleaned and rinsed with reagent water. The volume collected should be sufficient to
      insure a representative sample, allow for replicate analysis and laboratory fortified
      matrix analysis, if required, and minimize waste disposal.

 8.2   Special sampling requirements and precautions for chlorite.

      8.2.1   Sample bottles used for chlorite analysis must be opaque or amber to protect the
             sample from light

      8.2.2   When preparing the LFM, be aware that chlorite is an oxidant and may react
             with the natural organic matter in an untreated drinking water mafrix as a result
             of oxidative demand. If untreated water is collected for chlorite analysis, and
             subsequently used for the LFM, EDA preservation will not control this demand
             and reduced chlorite recoveries may be observed.

 8.3   Sample preservation and holding times for the anions are as follows:

	Analvte                 	Preservation            	Holding Time
      Bromate                        50 mg/L EDA, refrigerate at <6°C     28 days
      Chlorate                        50 mg/L EDA, refrigerate at <6°C     28 days
      Chlorite                         50 mg/L EDA, refrigerate at <6°C     14 days
      Bromide (source/raw water only)  EDA permitted, refrigerate at <6°C    28 days
      NOTE: Samples for chlorite analysis must arrive at the laboratory within 48 hours of
      collection and must be received at 10°C or less.

 8.4   When collecting a field sample from a treatment plant employing chlorine dioxide, the
      field sample must be sparged with an inert gas (helium, argon, nitrogen) prior to
      addition of the EDA preservative at time of sample collection.

 8.5   All four anions (bromate > 15.0 ug/L) can be analyzed by conductivity, in a sample
      matrix which has been preserved with EDA. Add a sufficient volume of the EDA
      preservation solution (Section 7.4) such that the final concentration is 50 mg/L in the
      sample. This would be equivalent to adding 0.5 mL of the EDA preservation solution
      to  1 L of sample.

 8.6   Chlorite is susceptible to degradation both through catalytic reactions with dissolved
      iron salts and reactivity towards free chlorine which exists as hypochlorous
      acid/hypochlorite ion in most drinking water as a residual disinfectant.11 EDA serves a
      dual purpose as a preservative for chlorite by chelating iron as well as any other
      catalytically destructive metal cations and removing hypochlorous acid/hypochlorite ion
      by forming an organochloramine.  EDA preservation  of chlorite also preserves the
      integrity of chlorate which can increase in unpreserved samples as a result of chlorite

                                     317.0-14

-------
        degradation.  EDA also preserves the integrity of bromate concentrations by binding
        with hypobromous acid/hypobromite ion which is an intermediate formed as a by-
        product of the reaction of either ozone or hypochlorous acid/hypochlorite ion with
        bromide ion.  If hypobromous acid/hypobromite ion is not removed from the matrix,
        further reactions may form bromate ion.

9.  QUALITY CONTROL

   9.1  Each laboratory using this method is required to operate a formal quality control (QC)
        program.  The requirements of this program consist of an initial demonstration of
        laboratory capability (IDC), and subsequent analysis in each analysis batch (Section
        3.1) of a Laboratory Reagent Blank (LRB), Initial Calibration Check Standard (ICCS),
        Laboratory Fortified Blank (LFB), Instrument Performance Check Standard (IPC),
        Continuing Calibration Check Standards (CCCS), Laboratory Fortified Sample Matrix
        (LFM) and either a Field, Laboratory or LFM duplicate sample analysis. This section
        details the specific requirements for each of these QC parameters for both the
        conductivity and absorbance detectors used in this application. Although this method
        involves both conductivity and absorbance detection, the MDLs and MRLs may differ
        but the QC requirements and acceptance criteria are the same for both detectors. The
        QC criteria discussed in the following sections are summarized in Section 17, Tables 4
        and 5. The laboratory is required to maintain performance records that define the
        quality of the data that are generated.
   9.2  INITIAL DEMONSTRATION OF CAPABILITY

        9.2.1   The Initial Demonstration of Capability (IDC) - This is used to characterize
               instrument and laboratory performance prior to performing analyses by this
               method. The QC requirements for the IDC discussed in the following section
               are summarized in Section 17, Table 4.

        9.2.2   Initial demonstration of low system background - Section 9.3.1.

        9.2.3   Initial Demonstration of Precision (IDP) - For the 4 conductivity detector
               analytes, prepare 7 replicate LFBs fortified at a recommended concentration of
               20 ug/L. For the absorbance detector, prepare 7 replicate LFBs fortified at a
               recommended concentration of 2.0 ug/L bromate. The percent relative standard
               deviation (RSD) of the results must be less than 20%.

        9.2.4   Initial Demonstration of Accuracy (IDA) - Using the data generated for Section
               9.2.3, calculate the average recovery. The average recovery of the replicate
               values must be within ± 15% of the true value.
                                       317.0-15

-------
     9.2.5   Quality Control Sample (QCS) - After calibration curves have initially been
            established or have been re-established, on a quarterly basis or as required to
            meet data quality needs, verify both the calibration and acceptable instrument
            performance with the preparation and analyses of an external/second source
            QCS. If the determined concentrations are not within ± 20% of the stated
            values, performance of the method is unacceptable. The source of the problem
            must be identified and corrected before proceeding with the IDC.

     9.2.6   Method Detection Limit (MDL) - MDLs must be established for all analytes,
            using reagent water (blank) fortified at a concentration of three to five times the
            estimated instrument detection limit.4 To determine MDL values, take seven
            replicate aliquots of the fortified reagent water and process through the entire
            analytical method. The replicates must be prepared and analyzed over three
            days. Report the concentration values in the appropriate units.  Calculate the
            MDL as follows:

                                 MDL = (t) x (S)

            where,  t =  student's t value for a 99% confidence level  and a standard
                        deviation estimate with n-1 degrees of freedom
                        [t = 3.14 for seven replicates], and
                    S = standard deviation of the replicate analyses.

            9.2.6.1  MDLs should be periodically verified, but MUST be initially
                    determined when a new operator begins work or whenever there is a
                    significant change in the background, or instrument response.

                    NOTE: Do not subtract blank values when performing MDL
                    calculations.

     9.2.7   Minimum Reporting Level (MRL) - The MRL is the threshold concentration of
            an analyte that a laboratory can expect to accurately quantitate in an unknown
            sample.  The MRL should be established at an analyte concentration either
            greater than three times the MDL or at a concentration which would yield a
            response greater than a signal to noise ratio of five. Setting the MRL too low
            may cause repeated QC failure upon analysis of the ICCS. Although the lowest
            calibration standard may be below the MRL,  the MRL must never be
            established at a concentration lower than the lowest calibration standard.

9.3  ASSESSING LABORATORY PERFORMANCE

     9.3.1   Laboratory Reagent Blank (LRB) - The laboratory must analyze at least one
            LRB with each analysis batch (Section 3.1). Data produced are used to assess
                                    317.0-16

-------
       contamination from the laboratory environment. Values that exceed Vz the MRL
       indicate a laboratory or reagent contamination is present. If a method analyte is
       observed in the LRB it must not exceed Vz the MRL. Analytes that exceed this
       level will invalidate the analysis batch for that method analyte in all
       corresponding field samples.

       9.3.1.1   EDA must be added to the LRB at 50 mg/L. By including EDA in the
                LRB, any bias as a consequence of the EDA which may be observed in
                the field samples, particularly in terms of background contamination,
                will be identified.

       9.3.1.2   When the PCR method is used for low level bromate analysis on
                samples from public water systems (PWSs) which employ chlorine
                dioxide disinfection, the matrix must be pretreated to remove the
                potentially interferant chlorite anion (Section 11.1.4). When these
                types of pretreated samples, or any type of pretreatment is applied to
                field samples included as  part of an analysis batch, a second LRB must
                be prepared, pretreated and analyzed to confirm no background effects
                of the pretreatenent are present. If the  analysis batch contains only
                pretreated samples, then only a pretreated LRB is required.

 9.3.2  Laboratory Fortified Blank (LFB) - Prepare a secondary dilution stock using the
       same stock solutions used to prepare the calibration standards and the LFM
       fortification solution. Since calibration solutions are prepared in large volumes
       and can be used over an extended period of time, the integrity of the
       concentration of the solution used to fortify the  LFM is checked by analyzing
       the LFB. The recovery of all analytes must fall in the acceptable recovery range,
       as indicated below, prior to analyzing samples.  If the LRB recovery for an
       analysis batch does not meet these recovery criteria the data are considered
       invalid, and the source of the problem must be identified and resolved before
       continuing with analyses.

	LFB Fortified Concentration range	LFB Percent Recovery Limits
	 MRL to 5 x MRL                                 75 - 125 %
	   5 x MRL to  highest calibration level                85 - 115 %

       9.3.2.1   EDA must be added to the LFB at 50 mg/L. The addition of EDA to
                all reagent water prepared calibration and quality control samples is
                required not as a preservative but rather as a means to normalize any
                bias attributed by the presence of EDA in the field samples.

 9.3.3  Instrument Performance Check (IPC) - The Initial Calibration Check Standard
       (ICCS) is to be evaluated as the IPC solution in order to confirm proper


                                317.0- 17

-------
instrument performance. As specified in Section 10.3.1, this must be done using
the lowest calibration standard or the standard level established as the MRL.
This analysis confirms the MRL and demonstrates proper chromatographic
performance at the beginning of each analysis batch.  Chromatographic
performance is judged by calculating the Peak Gaussian Factor (PGF), which is
a means to measure peak symmetry and monitoring retention time drift in the
surrogate peak over time. If these criteria are not met, corrective action must be
performed prior to analyzing additional samples. Major maintenance like
replacing columns require rerunning the IDC (Section 9.2).

9.3.3. 1  Critically evaluate the surrogate peak in the initial calibration check
        standard, and calculate the PGF as follows:

                        1.83 x WO/2)
                 PGF = -----------------------
                          W(V10)
        where,   W(1A) is the peak width at half height, and
                 W (V10) is the peak width at tenth height.

        NOTE: Values for W(x/2) and W (V10) can be attained through most
        data acquisition software.

9.3.3.2  Small variations in retention time can be anticipated when a new
        solution of eluent is prepared but if sudden shifts of more than 5% are
        observed in the surrogate retention time, some type of instrument
        problem is present. Potential problems include improperly prepared
        eluent, erroneous method parameters programmed such as flow rate or
        some other system problem. The chromatographic profile (elution
        order) of the target anions following an ion chromatographic analysis
        should closely replicate the profile displayed in the test chromatogram
        that was shipped when the column was purchased. As a column ages, it
        is normal to see a gradual shift and shortening of retention times, but if
        after several years of use, extensive use over less than a year, or use
        with harsh samples, this retention time has noticeably shifted to any
        less than 80% of the original recorded value, the column requires
        cleaning or replacement; especially if resolution problems are
        beginning to become common between previously resolved peaks. A
        laboratory should retain a historic record of retention times for the
        surrogate and all the target anions to provide evidence of an analytical
        columns vitality.
                        317.0-18

-------
            9.3.3.3  If a laboratory chooses to monitor exclusively for trace bromate using
                    PCR and the UV/VIS absorbance detector, and no other analytes are
                    being monitored on the conductivity detector, the surrogate maybe
                    omitted from the procedure.  In this case, no measurement of PGF is
                    required.  However,  the laboratory must carefully monitor the bromate
                    retention time in the ICCS as an alternate to the surrogate retention
                    time and, in the same manner, adhere to those specifications set forth
                    in Section 9.3.3.2.  During the course of the analysis, bromate retention
                    times in the CCCS and ECCS must also be closely monitored to be
                    certain they adhere to the QC requirements set forth in Section
                    10.3.2.2.

9.4  ASSESSING ANALYTE RECOVERY AND DATA QUALITY

     9.4.1   Laboratory Fortified Sample Matrix (LFM) - The laboratory must add a known
            amount of each target analyte to a minimum of 5% of the collected field samples
            or at least one with every analysis batch,  whichever is greater.  Additional LFM
            requirements, as described in Section 9.4.1.5, apply when the PCR system is
            used for low level bromate in chlorine dioxide disinfected waters. For a LFM to
            be valid, the target analyte concentrations must be greater than the native level
            and must adhere to the requirement outlined in Section  9.4.1.2. It is
            recommended that the solutions used to fortify the LFM be prepared from the
            same stocks used to prepare the calibration standards and not from external
            source stocks.  This will remove the bias contributed by an externally prepared
            stock and focus on any potential bias introduced by the field sample matrix.

            9.4.1.1  The fortified concentration must be equal to or greater than the native
                    concentration. Fortified samples that exceed the calibration range
                    must be diluted to be within the linear range. In the event that the
                    fortified level is less than the observed native level of the unfortified
                    matrix, the recovery should not be calculated.  This is due to the
                    difficulty in calculating accurate recoveries of the fortified
                    concentration when the native sample concentration to fortified
                    concentration ratio is greater than one.

            9.4.1.2  The LFM should be prepared at concentrations no greater than ten
                    times the highest concentration observed in any field sample and
                    should be varied to reflect the range of concentrations observed infield
                    samples. If no analytes are observed in any field sample, the LFM
                    should be fortified near the MRL.
                                    317.0-19

-------
       9.4.1.3  Calculate the percent recovery for each target analyte, corrected for
               concentrations measured in the unfortified sample. Percent recovery
               should be calculated using the following equation:

                                  (C. - C)
                        %REC =	 x 100
               where,   %REC =  percent recovery,
                        Cs =  fortified sample concentration,
                        C = native sample concentration, and
                        s  =  concentration equivalent of analyte added to sample.

       9.4.1.4  Recoveries may exhibit a matrix dependence. If the recovery of any
               analyte falls outside 75 - 125%, and the laboratory's performance for
               all other QC performance criteria are acceptable, the accuracy problem
               encountered with the fortified sample is judged to be matrix related,
               not system related. The result for that analyte in the unfortified sample
               and the LFM must be labeled suspect/matrix to inform the data user
               that the result is suspect due to matrix effects. Repeated failure to
               meet suggested recovery criteria indicates potential problems with the
               procedure and should be investigated.

       9.4.1.5  When the PCR method is used for low level bromate analysis on field
               samples from PWSs which employ chlorine dioxide disinfection and
               consequently contain chlorite, a LFM must be prepared, exclusively
               for trace bromate, for each of these field samples. Initially, the field
               sample is analyzed and chlorite, chlorate and bromide levels are
               determined. Then, a second aliquot of field sample ispretreated to
               remove chlorite, as described in Section 11.1.4, and analyzed to
               determine native bromate concentration. A third aliquot of the field
               sample then must be fortified with bromate, pretreated as described in
               Section 11.1.4 to remove chlorite, and analyzed to assess bromate
               recovery from that matrix.  This additional QC is required to rule out
               matrix effects  and to confirm that the laboratory performed the chlorite
               removal step (Section 11.1.4.1) appropriately. This LFM should be
               fortified with bromate at concentrations close to but greater than the
               level determined in the native sample. Recoveries are determined as
               described above (Section 9.4.1.3).  Samples that fail the LFM percent
               recovery criteria of 75 -125% must be reported as suspect/matrix.

9.4.2   SURROGATE RECOVERY- The surrogate is specific to the conductivity
       detector and shows no response on the postcolumn absorbance detector.
                               317.0-20

-------
       Calculate the surrogate recovery for the conductivity detector from all analyses
       using the following formula:

                                   SRC
                         %REC =	 xlOO
                                   SFC

       where, %REC = percent recovery,
               SRC = surrogate recovered concentration, and
               SFC = surrogate fortified concentration.

       9.4.2.1  Surrogate recoveries must fall between 90-115% for proper instrument
               performance and analyst technique to be verified.  The recovery range
               for the surrogate is extended to 115% to allow for the potential
               contribution of trace levels of dichloroacetate as a halogenated organic
               disinfection by-product (DBF) of dichloroacetic acid (DCAA).
               Background levels of this organic DBF are rarely observed above 50
               l-ig/L (0.05 mg/L) which constitutes only 5% of the 1.00 mg/L
               recommended fortified concentration.

       9.4.2.2  If the surrogate recovery falls outside the 90-115% recovery window,
               an analysis error is evident and sample reanalysis is required.  Poor
               recoveries could be the result of imprecise sample injection or analyst
               fortification errors.  If the second analysis also fails the recovery
               criterion, report all data for that sample as suspect.

       9.4.2.3  If a laboratory chooses to monitor exclusively for trace bromate using
               PCR and the UV/VIS absorbance detector, and no other analytes are
               being monitored on the conductivity detector, the surrogate maybe
               omitted from the procedure.  In this situation, the laboratory MUST
               adopt the QC protocol outlined in Section 9.3.3.3.

9.4.3   FIELD OR LABORATORY DUPLICATES - The laboratory must analyze
       either a field or a laboratory duplicate for a minimum of 5% of the collected
       field samples or at least one with every analysis batch, whichever is greater.
       The sample matrix selected for this duplicate analysis must contain measurable
       concentrations of the target anions in order to establish the precision of the
       analysis set and insure the quality of the data. If none of the samples within an
       analysis batch have measurable concentrations, the LFM should be repeated as a
       laboratory duplicate.

       9.4.3.1  Calculate the relative percent difference (RPD) from the mean using
               the following formula:
                               317.0-21

-------
                                 RPD =	xlOO
                                       (Pc + Dc]/2)

                        where,   RPD = relative percent difference
                                 Ic = the initial quantitated concentration, and
                                 Dc =  the duplicate quantitated concentration

               9.4.3.2   Duplicate analysis acceptance criteria.
                        Concentration range	RPD Limits
                        MRL to 5 x MRL                      ± 20 %
                        5 x MRL to highest calibration level     ± 10 %

               9.4.3.3   If the RPD  for any target analyte falls outside the acceptance criteria
                        (Section 9.4.3.2) and if all other QC performance criteria are met for
                        that analyte, the result for the sample and duplicate should be labeled
                        as suspect/matrix to inform the data user that the result is suspect due
                        to a potential matrix effect, which led to poor precision This should
                        not be a chronic problem and if it frequently recurs (>20% of duplicate
                        analyses), it indicates a problem with the instrument or analyst
                        technique that must be corrected.

        9.4.4  In recognition of the rapid advances occurring in chromatography, the analyst is
               permitted certain options, such as the use of different columns, injection
               volumes, and/or eluents, to improve the separations or lower the cost of
               measurements. Each time such modifications to the method are made, the
               analyst is required to repeat the procedure in Section 9.2 and adhere to the
               condition of conductivity baseline stability found in Section 1.2.1.

        9.4.5  It is recommended that the laboratory adopt additional quality assurance (QA)
               practices for use with this method. The specific practices that are most
               productive depend upon the needs of the laboratory and the nature of the
               samples. Whenever possible, the laboratory  should perform analysis of quality
               control check standards and participate in relevant proficiency testing (PT) or
               performance evaluation (PE) sample studies.

10. CALIBRATION AND STANDARDIZATION

   10.1 Demonstration and documentation of acceptable initial calibration is required prior to
        the IDC  and before any samples are analyzed, is required intermittently throughout
        sample analysis to meet required QC performance criteria outlined in this method and is
        summarized in Tables 4 and 5. Initial calibration verification is performed using a QCS
        as well as with each analysis batch using an initial, continuing (when more than 10 field


                                        317.0-22

-------
     samples are analyzed), and end calibration standards.  The procedures for establishing
     the initial calibration curve are described in Section 10.2. The procedures to verify the
     calibration with each analysis batch is described in Section 10.3.

10.2 INITIAL CALIBRATION CURVE

     10.2.1  Establish ion chromatographic operating parameters equivalent to those
            indicated in Table 1 and configured as shown in Figure 1.

     10.2.2  Estimate the Linear Calibration Range - The linear concentration range is the
            concentration range over which the instrument response is linear. On the
            conductivity detector for the four target analytes (chlorite, bromate, bromide and
            chlorate) the linear range should cover the expected concentration range of the
            field samples and should not extend over more than two orders of magnitude in
            concentration. The restriction of two  orders of magnitude is prescribed since
            beyond this it is difficult to maintain linearity throughout the entire calibration
            range.

            10.2.2.1 If quantification is desired over a larger range, then two separate
                    calibration curves must be prepared.

            10.2.2.2 For an individual calibration curve, a minimum of three calibration
                    standards are required for a curve that extends over a single order of
                    magnitude and a minimum of five calibration standards are required if
                    the curve covers two orders of magnitude. Because high
                    concentrations of chlorite can interfere with the postcolumn analysis of
                    low levels of bromate, the conductivity and absorbance detectors must
                    be calibrated separately.

            10.2.2.3 Since the concentration ranges in actual field samples by conductivity
                    detection for chlorite, bromide and chlorate are expected to cover two
                    orders of magnitude, the use of at least five calibration standards in the
                    range 5 - 500 |-ig/L is recommended. Bromate concentrations are
                    expected to  be significantly lower. It is suggested that the conductivity
                    detector be calibrated using at least five bromate calibration standard
                    levels in the range 5-100 |-ig/L. Additionally, report values for
                    bromate by  conductivity ONLY when they are measured by the PCR
                    above  15.0 ug/L. The conductivity detector will observe a response for
                    bromate at concentration below 15.0 ug/L but concentrations between
                    5.0 and 15.0 ug/L are within the calibrated range for PCR detection
                    and will reflect far better precision and accuracy.

            10.2.2.4 Although the bromate calibration  curve for the absorbance detector


                                    317.0-23

-------
               extends over less than two orders of magnitude, the use of five
               calibration standards, containing only bromate in the range 0.5-15.0
               l-ig/L, is recommended.

10.2.3  Prepare the calibration standards by carefully adding measured volumes of one
       or more stock standards (Section 7.3) to a volumetric flask and diluting to
       volume with reagent water.  Prior to using mixed standards for calibration, it
       must be ensured that the individual calibration standards do not contain any
       appreciable concentrations of the other target analytes.

       10.2.3.1  EDA must be added to the calibration standards at 50 mg/L. The
               addition of EDA to all reagent water prepared calibration and quality
               control samples is required not as a preservative but rather as a means
               to normalize any bias contributed by the addition of EDA to preserve
               the field samples.

       10.2.3.2 Prepare a 10.0 mL aliquot of surrogate fortified calibration solution
               which can be held for direct manual injection or used to fill an
               autosampler vial. This is done by adding 20 |_iL of the surrogate
               solution (Section 7.5) to a 20 mL disposable plastic micro beaker.
               Next, transfer 10.0 mL of calibration standard into the micro beaker
               and mix.  These  volumes may be adjusted to meet specific laboratory
               autosampler volume requirements provided the fortified surrogate
               concentration is  at the prescribed concentration of 1.0 mg/L.  The
               calibration standard is now ready for analysis. The same surrogate
               solution that has been employed for the standards should also be used
               in Section 11.1 for the field samples.

               NOTE: This surrogate fortification procedure may be omitted if a
               laboratory chooses to monitor exclusively for trace bromate using PCR
               and the UV/VIS absorbance detector, and no other analytes are being
               monitored on the conductivity detector.  In this situation, the
               laboratory must adopt the QC protocol outlined in Section 9.3.3.3.

10.2.4  Inject 225  |_iL of each calibration standard. Increased sensitivity for low level
       detection of bromate by PCR can be  achieved by increasing the injected sample
       volume.4  If the injection volume is increased special operating conditions must
       be used to  insure proper chromatographic performance.4

10.2.5  Tabulate peak area responses against the concentration for the four target
       analytes, the surrogate from the conductivity detector and bromate from the
       postcolumn absorbance detector.   The results are used to prepare calibration
       curves using linear regression analysis for each analyte on the conductivity
       detector and using a quadratic regression analysis for bromate on the absorbance

                               317.0-24

-------
            detector.

            10.2.5.1  Use of peak areas are strongly recommended since they have been
                    found to be more consistent, in terms of quantitation, than peak
                    heights.  Peak height can tend to be suppressed as a result of high
                    levels of common anions in a given matrix which can compete for
                    exchange sites leading to peak broadening. Using peak areas, it is the
                    analyst responsibility to review all chromatograms to insure accurate
                    baseline integration of target analyte peaks, since poorly drawn
                    baselines will more significantly influence peak areas than peak
                    heights.

     10.2.6  After establishing or reestablishing calibration curves, the accuracy of this
            calibration must be verified through the analysis of a QCS or an externally
            prepared second source standard. The QCS should be prepared at a
            concentration near the middle of the calibration and is best to be analyzed in
            triplicate. As specified in Section 9.2.5, determined concentrations must fall
            within ±15% of the stated values.

10.3  CONTINUING CALIBRATION VERIFICATION - Initial calibrations maybe stable
     for extended periods of time. Once the calibration curves have been established for
     both the conductivity and absorbance detectors, they must be verified for each analysis
     batch, prior to conducting any field sample analyses using an Initial Calibration Check
     Standard. Continuing Calibration Check Standards and End Calibration Check
     Standards are also required as described in the sections below.
     10.3.1  INITIAL CALIBRATION CHECK STANDARD (ICCS) - The initial
            calibration must be determined to be valid each day prior to analyzing any
            samples. Since two detectors are incorporated in this method,  this must be
            accomplished by using a mixed calibration check standard for the four
            conductivity analytes and a separate low level bromate initial calibration check
            standard for the absorbance detector.  In both cases, the lowest level standard
            used to prepare the calibration curve must be used. In cases where the analyst
            has chosen to set the MRL above the lowest standard, a standard at a
            concentration equal to or below the MRL is acceptable. Percent recovery for the
            ICCS must be in the range or 75 - 125% before the analyst is allowed to analyze
            samples.

     10.3.2  CONTINUING CALIBRATION CHECK/END CALIBRATION CHECK
            STANDARDS (CCCS/ECCS) - Continuing calibration check standards must
            be analyzed after every tenth field sample analysis and at the end of the analysis
            batch as an end calibration check standard.  For the reasons noted above, two
            separate continuing and end calibration check standards must be incorporated.
            If more than 10 field samples are included in an analysis batch, the analyst must
                                   317.0-25

-------
               alternate between the middle and high continuing calibration check standard
               levels.

               10.3.2.1  The percent recovery for the CCCS/ECCS must meet the following
                        criteria:

	Concentration range	Percent Recovery Limits
       	          MRL to 5 x MRL                     75 - 125 %
     	            5 x MRL to highest calibration level    85 - 115 %

               10.3.2.2  If during the analysis batch, the measured concentration on either
                        detector differs by more than the calibration verification criteria shown
                        above, or the retention times shift more than ± 2% from the last
                        acceptable initial or continuing calibration check standard for any
                        analyte, all samples analyzed after the last acceptable calibration check
                        standard are considered invalid and must be reanalyzed. The source of
                        the problem must be identified and resolved before reanalyzing the
                        samples or continuing with the analyses.

               10.3.2.3  In the case where the end calibration failed to meet performance
                        criteria, but the initial and middle calibration check standards were
                        acceptable, the samples bracketed by the acceptable calibration check
                        standards may be reported. However, all field samples between the
                        middle and end calibration check standards must be reanalyzed.

11. PROCEDURE

   11.1  SAMPLE PREPARATION

         11.1.1 For refrigerated or field samples arriving to the laboratory cold, ensure the
               samples have come to room temperature prior to conducting sample analysis by
               allowing the samples to warm on the bench for at least 1 hour.

         11.1.2 Prepare a 10.0 mL aliquot of surrogate fortified sample which can beheld for
               direct manual injection or used to fill an autosampler vial. This is done by
               adding 20 |_iL of the surrogate solution (Section 7.5) to a 20 mL disposable
               plastic micro beaker. Next, place a 10.0 mL aliquot of sample into  the micro
               beaker and mix. These volumes may be adjusted to meet specific laboratory
               autosampler volume requirements provided the fortified surrogate concentration
               is at the prescribed concentration of 1.0 mg/L. The sample is now ready for
               analysis.

               NOTE:  The less than 1% dilution error introduced by the addition of the


                                        317.0-26

-------
       surrogate is considered insignificant. In addition, this surrogate fortification
       procedure may be omitted if a laboratory chooses to monitor exclusively for
       trace bromate using PCR and the UV/VIS absorbance detector, and no other
       analytes are being monitored on the conductivity detector.  In this situation, the
       laboratory must adopt the QC protocol outlined in Section 9.3.3.3.

11.1.3  Using a Luer lock, plastic 10 mL syringe, withdraw the sample from the micro
       beaker and attach a 0.45 |_im particulate filter (demonstrated to be free of ionic
       contaminants) directly to the syringe. Filter the sample into an autosampler vial
       (if vial is not designed to automatically filter) or manually load the injection
       loop injecting a fixed amount of filtered, well mixed sample. If using a
       manually loaded injection loop, flush the loop thoroughly between sample
       analysis using sufficient volumes of each new sample matrix.

11.1.4  CHLORINE DIOXIDE - TREATED WATERS CONTAINING CHLORITE -
       Treatment plants that use chlorine dioxide as part of their treatment process can
       produce high levels of chlorite in samples. Since chlorite can interfere with the
       postcolumn quantitation of low levels of bromate as described in Section 4.6,
       chlorite must be removed from these samples prior to analysis.12 The oxidation-
       reduction reaction between ferrous iron and chlorite13 is used to remove chlorite
       without any adverse  affects on the bromate concentration.14 The EDA stabilized
       sample is acidified to a pH of 5-6 (verified using pH test strips), ferrous iron
       solution is added and allowed to react for 10 minutes.  The sample is then
       filtered using  a 0.45 micron membrane to remove precipitated ferric hydroxide
       and the excess soluble  iron is removed by passing the filtered sample through a
       hydrogen cartridge [a solid phase extraction (SPE)  clean-up cartridge in the H+
       form, (Section 6.11)], prior to analysis. Prior to using any pretreatment, each lot
       of cartridges must be QC checked to insure proper  analyte recoveries are
       maintained and laboratory reagent blanks are free from interferences. In
       addition, consistent lots of reagents, pretreatment cartridges, and membrane
       cartridges must be used throughout an entire analysis batch to maintain assured
       QC uniformity.

       11.1.4.1 Place a 10 mL aliquot of sample in a 20 mL micro beaker and add 35
               uL of 0.5 N sulfUric acid (Section 7.8).  After mixing, verify thepH is
               between 5 and 6 using pH test strips, add  40 uL of ferrous iron solution
               (Section 7.7), mix and allow to react for 10 minutes. Filter the
               reaction mixture using a 0.45 micron particulate filter (Section 6.10)
               attached to a  10 mL syringe into the  barrel of a second syringe to
               which a pre-conditioned hydrogen cartridge (Section 6.11) is attached.
               Pass the solution through a hydrogen cartridge  at  a flow rate of
               approximately 2 mL per minute.  Discard the first 3 mL, and collect an
               appropriate volume (depending on autosampler vial size) for analysis.
                               317.0-27

-------
                    Add the respective volume of surrogate solution, depending on the
                    volume collected, and the sample is ready for analysis.

                    NOTE: Pretreated samples can be held for no more than 30 hours after
                    initial pretreatment.  If this time has expired, the pretreatment steps
                    must be repeated on a second aliquot of both the field sample matrix
                    and the respective LFM.

            11.1.4.2 In order to ensure data quality, all samples from PWSs which utilize
                    chlorine dioxide which have been pretreated to remove chlorite,
                    MUST also be used to prepare a pretreated LFM specific to trace
                    bromate. This LFM should be fortified with bromate at concentrations
                    close to but greater than the level determined  in the native sample.
                    Initially, the field sample is analyzed and chlorite, chlorate and
                    bromide levels are determined. Then, a second aliquot of field sample
                    is pretreated to remove chlorite, as described above and analyzed to
                    determine native bromate concentrations.  A third aliquot of the field
                    sample then must be fortified with bromate, pretreated to remove
                    chlorite, and analyzed to assess bromate recovery from that matrix.
                    This additional QC is required to rule out matrix effects and to  confirm
                    that the laboratory performed the chlorite removal step appropriately.
                    If the bromate recovery falls outside the acceptance range of 75 - 125%
                    (Section 9.4.1.5), that particular sample should be reported as
                    suspect/matrix.

            11.1.4.3 Suppressor devices which have had long term exposure to iron cations
                    may have reduced method performance in other applications, such as
                    the determination of certain common inorganic anions.  If reduced
                    peak response is observed, particularly for fluoride or phosphate, the
                    ASRS should be cleaned according to the manufacturer's
                    recommendations.

11.2 SAMPLE ANALYSIS

     11.2.1  Table 1 summarizes the recommended operating conditions for the ion
            chromatograph and delivery of the postcolumn reagent. Included in this table is
            estimated retention times that can be achieved by this method.  Other columns
            or chromatographic conditions may be used if the requirements of Section 9.2
            are met.

     11.2.2  Establish a valid initial calibration as described in Section 10.2 and complete the
            IDC (Section  9.2). Check system calibration by analyzing an ICCS (Section
            10.3.1) as part of the initial QC for the analysis batch and, if required,
            recalibrate as  described in Section 10.3.

                                    317.0-28

-------
     11.2.3  Inject 225 |_iL of each sample. Use the same size loop for standards and
            samples. An automated constant volume injection system may also be used.

            11.2.3.1  Increased sensitivity for low level detection of bromate by PCR can be
                    achieved by increasing the injected sample volume.4 If the injection
                    volume is increased (Section 10.2.4) special operating conditions must
                    be used to insure proper chromatographic performance.4

     11.2.4  The width of the retention time window used to make identifications should be
            based upon measurements of actual retention time variations of standards
            measured over several days. Three times the  standard deviation of retention
            time can be used to calculate a suggested window size for each analyte.
            However, the experience of the  analyst should weigh heavily in the
            interpretation of chromatograms.

     11.2.5  If the response of a sample analyte exceeds the calibration range,  the sample
            must be diluted with an appropriate amount of EDA fortified reagent water and
            reanalyzed. If this is not possible then three new calibration concentrations
            must be employed to create a separate high concentration calibration curve, one
            standard near the estimated concentration and the other two bracketing around
            an interval equivalent to approximately ± 25% the estimated concentration. The
            latter procedure involves significantly more time than a simple sample dilution
            and, therefore, it is advisable to collect sufficient sample to allow for sample
            dilution and sample reanalysis, if required.
     11.2.6  Should more complete resolution be needed between any two coeluting peaks,
            the eluent (Section 7.2) can be diluted. This will spread out the run, however,
            and will cause late eluting anions to be retained even longer. The analyst must
            verify that this dilution does not negatively affect performance by repeating and
            passing all the QC criteria in Section 9, and by reestablishing a valid initial
            calibration curve (Section 10.2). As a specific precaution, upon dilution of the
            carbonate eluent, a peak for bicarbonate maybe observed by conductivity within
            the retention time window for bromate which will negatively impact the
            analysis.

            11.2.6.1  Eluent dilution will reduce the overall response of an anion due to
                    chromatographic band broadening which will be evident by shortened
                    and broadened peaks.  This will adversely effect the MDLs for each
                    analyte.

11.3  AUTOMATED ANALYSIS WITH METHOD 317.0

     11.3.1  Laboratories conducting analyses on large numbers of samples often prepare


                                    317.0-29

-------
               large analysis batches that are run in an automated manner. When conducting
               automated analyses, careful attention must be paid to all reservoirs to be certain
               sufficient volumes are available to sustain extended operation. Laboratories
               must ensure that all QC performance criteria are met as described in preceding
               sections to ensure their data are of acceptable quality.

               11.3.1.1  Special attention must be made when the PCR reservoir is refilled.
                        Since this is a pneumatically driven system, the baseline will require a
                        minimum often minutes to restabilize after the reservoir has been
                        refilled and the bottle repressurized.

         11.3.2 Because this method has two detectors that require independent calibration,
               analysis sequences must be carefully constructed to meet required QC
               specifications and frequency (Table 5). To help with this task, an acceptable
               sequence for a sample analysis batch, with all the method-required QC, is shown
               in Table 6. This schedule is included only as an example of a hypothetical
               analysis batch where the analyst desires to collect data using both detectors.
               Within the analysis batch, references to exact concentrations for the ICCS,
               CCCS and ECCS are for illustrative purposes only. The analyses for sample
               #14 provides an example of the QC requirements for a complete conductivity
               and trace bromate PCR analysis of a sample from a PWS employing chlorine
               dioxide disinfection.

         11.3.3 Table 6 may be used as a guide when preparing analysis batches.

12. DATA ANALYSIS AND CALCULATIONS

   12.1  Identify the method analytes in the sample chromatogram by comparing the retention
         time of the suspected analyte peak to the retention time  of a known analyte peak in a
         calibration standard. If analyte retention times have shifted (generally towards shorter
         times) since the initial calibration, but are still within acceptance criteria and are
         reproducible during the analysis batch, the analyst should use the retention time in the
         daily calibrations to confirm the presence or absence of target analytes.

   12.2  Compute sample concentration using the initial calibration curve generated in Section
         10.2.

   12.3  Report ONLY those values that fall between the MRL and the highest calibration
         standards. Samples with target analyte responses exceeding the highest standard must
         be diluted and reanalyzed. When this is not possible the alternate calibration
         procedures described in Section 11.2.5 must be followed.  Samples with target analytes
         identified but  quantitated below the concentration established by the lowest calibration
         standard may be reported as present, but below the minimum reporting limit (MRL),
         and consequently not quantitated.

                                        317.0-30

-------
         12.3.1 Report bromate concentrations using the postcolumn UV/VIS absorbance
               detector when they fall between the MRL and 15.0 ug/L. When bromate
               concentrations exceed 15.0 ug/L, as detected by UV/VIS absorbance, either
               report by conductivity, calibrate the postcolumn UV/VIS absorbance detector to
               a higher bromate concentration, or dilute the sample.

   12.4  Report results in |_ig/L.

   12.5  Software filtering of the postcolumn UV/VIS absorbance signal is required to improve
         the precision of peak measurements, minimize non-random noise and improve peak
         appearance. Olympic smoothing (25 points, 5 seconds with 1 iteration) was chosen
         using peak area for quantitation because it was determined to have minimal effect on
         peak height and/or area.2'15 The use of alternate smoothing routines is acceptable
         providing all QC criteria are met.

13. METHOD PERFORMANCE

   13.1  Table 1  gives the standard conditions, typical retention times and single laboratory
         MDLs in reagent water, as determined for each of the inorganic oxyhalide DBFs and
         bromide.  Included in this table is a comparison of the MDLs determined by
         conductivity both with and without the postcolumn UV/VIS absorbance system on-line.
         These data indicate that the postcolumn UV/VIS detector system has no effect on
         conductivity detector performance (carefUl attention must however be paid to insure
         backpressure on the suppressor is kept below 120 psi).
   13.2  Table 2 shows the precision and accuracy of the trace bromate measurement, evaluated
         on both detectors, at two fortified concentrations, in chlorinated surface water, a
         simulated high ionic strength water (HIW) and a simulated high organic (HOW) content
         water. The mean bromate recovered concentration (accuracy relative to the fortified
         level) and the precision (expressed as %RSD of the replicate analyses) are tabulated.
         The HIW was designed to simulate a high ionic strength field sample and the HOW
         designed to simulate a high organic content field sample.  The HIW was prepared from
         reagent water which was fortified with the common anions of chloride at  100 mg/L,
         carbonate at 100 mg/L, nitrate at 10.0 mg/L as nitrogen, phosphate at 10.0 mg/L as
         phosphorous, and sulfate at 100 mg/L.1  The HOW was prepared from reagent water
         fortified with 1.0 mg/L fulvic acid.1

   13.3  Table 3 gives the single laboratory standard deviation and precision (% RSD) for each
         anion included in the method in a variety of waters for the standard conditions
         identified in Table I.1'2

   13.4  Table 3 A shows the stability data for the inorganic oxyhalide DBFs. Each data point in
         these tables represent the mean percent recovery following triplicate analyses.  These

                                       317.0-31

-------
        data were used to formulate the holding times shown in Section 8.3.1

14. POLLUTION PREVENTION

   14.1 Pollution prevention encompasses any technique that reduces or eliminates the quantity
        or toxicity of waste at the point of generation. Numerous opportunities for pollution
        prevention exist in laboratory operation. The EPA has established a preferred hierarchy
        of environmental management techniques that places pollution prevention as the
        management option of first choice. Whenever feasible, laboratory personnel should use
        pollution prevention techniques to address their waste generation. When wastes cannot
        be feasiblely reduced at the source, the Agency recommends recycling as the next best
        option.

   14.2 Quantity of the chemicals purchased should be based on expected usage during its
        shelf-life and disposal cost of unused material. Actual reagent preparation volumes
        should reflect anticipated usage and reagent stability.

   14.3 For information about pollution prevention that may be applicable to  laboratories and
        research institutions, consult "Less is Better: Laboratory Chemical Management for
        Waste Reduction," available from the American Chemical Society's Department of
        Government Regulations and Science Policy, 1155 16th Street N.W., Washington B.C.
        20036, (202) 872-4477.
15. WASTE MANAGEMENT

   15.1 The Environmental Protection Agency requires that laboratory waste management
        practices be conducted consistent with all applicable rules and regulations. Excess
        reagents, samples and method process wastes should be characterized and disposed of
        in an acceptable manner.  The Agency urges laboratories to protect the air, water, and
        land by minimizing and controlling all releases from hoods and bench operations,
        complying with the letter and spirit of any waste discharge permit and regulations, and
        by complying with all solid and hazardous waste regulations, particularly the hazardous
        waste identification rules and land disposal restrictions. For further information on
        waste management consult the "Waste Management Manual for Laboratory Personnel,"
        available from the American Chemical Society at the address listed in Section 14.3.

16. REFERENCES

1.   U.S. EPA Method 300.1. EPA Document number: EPA/600/R-98/118. NTIS number
    PB98-169196INZ.
                                       317.0-32

-------
2.   Wagner, H.P., Pepich, B.V., Hautman, D.P. and Munch, D.J. "Analysis of 500 ng/L Levels
    of Bromate in Drinking Water by Direct Injection and Suppressed Ion Chromatography
    Coupled with a Single, Pneumatically Delivered Postcolumn Reagent."  Journal
    ChromatographvA. 850 (1999), 119-129.

3.   Wagner, H.P., Alig, A.E., Pepich, B.V., Frebis, C.P., Hautman, D.P. and Munch, D.J. "A
    Study of Ion Chromatographic Methods for Trace Level Bromate Analysis in Drinking
    Water Comparing the Selective Anion Concentration (SAC) Method, U.S. EPA 300.1 and a
    Postcolumn Reagent Procedure." AWWA WOTC Proceedings. 4D-1, San Diego, CA,
    (November 1998).

4.   Wagner, H.P., Pepich, B.V., Hautman, D.P. and Munch, D.J. "Performance Evaluation of a
    Method for the Determination of Bromate in Drinking Water by Ion Chromatography (EPA
    317.0) and validation of EPA Method 324.0 " Presented at the International Ion
    Chromatography Symposium, San Jose, CA, September, 1999. Accepted for publication
    Journal of ChromatographvA. (anticipated) June 2000.

5.   Glaser, J.A., Foerst, D.L., McKee, G.D., Quave, S.A., and Budde, W.L.  "Trace Analyses
    for Wastewater," Environmental Science and Technology. Vol.  15, Number 12, page 1426,
    December, 1981.

6.   "OSHA Safety and Health Standards, General Industry," (29CFR1910). Occupational Safety
    and Health Administration, OSHA 2206, (Revised, Jan. 1976).
7.   ASTM Annual Book of Standards, Part II, Volume 11.01, D3370-82, "Standard Practice for
    Sampling Water," American Society for Testing and Materials, Philadelphia, PA, 1986.

8.   "Carcinogens-Working with Carcinogens," Publication No. 77-206, Department of Health,
    Education, and Welfare, Public Health Service, Center for Disease Control, National
    Institute of Occupational Safety and Health, Atlanta, Georgia, August 1977.

9.   "Safety In Academic Chemistry Laboratories," 3rd Edition, American Chemical Society
    Publication, Committee on Chemical Safety, Washington, D.C., 1979.

10. Standard Methods for the Examination of Water and Wastewater, "Method 4500-C1O 2,C
    Amperometric Method I (for the determination of Chlorine Dioxide)," 19th Edition of
    Standard Methods (1995).

11. Hautman, D.P. & Bolyard, M.  "Analysis of Oxyhalide Disinfection By-products and other
    Anions of Interest in Drinking Water by Ion Chromatography." Journal of Chromatographv.
    602, (1992 ), 65-74.
                                      317.0-33

-------
12.  Wagner, H.P., Pepich, B.V., Hautman, D.P. and Munch, D.J. "The Use of EPA Method
    300.1 with the Addition of a Postcolumn Reagent for the Analysis of Ultra Trace Levels of
    Bromate in Drinking Water and the Analysis of Bromate in Bottled Waters." Presented at
    Pittcon 99, Orlando FL, paper 1414, (March 1999).

13.  latrou, A. and Knocke, W.R. Removing Chlorite by the Addition of Ferrous Iron. Journal of
    the AWWA Research and Technology, (November, 1992), 63-68.

14.  Wagner, H.P., Pepich, B.V., Hautman, D.P. and Munch, D.J. "Eliminating the Chlorite
    Interference in US Environmental Protection Agency Method 317.0 Permits the Analysis of
    Bromate in all Drinking Water Matrices."  Accepted for publication in Journal of
    Chromatography A. (anticipated) Fall 2000.

15.  Schibler, J.A. American Laboratory. (December, 1997), 63-64.

16.  Dixon, W.J. "Processing Data Outliers." Biometrics. BIOMA 9 (No.l):74-89 (1953).
                                      317.0-34

-------
17. TABLES. DIAGRAMS. FLOWCHARTS AND VALIDATION DATA

TABLE 1.  CHROMATOGRAPHIC CONDITIONS AND METHOD DETECTION
           LIMITS IN REAGENT WATER FOR THE INORGANIC OXYHALIDE
           DISINFECTION BY-PRODUCTS AND BROMIDE.
Standard Conditions and Equipment*
                                         a).
Ion Chromatograph:
Sample Loop:
Eluent:
Eluent Flow:
Columns :
Typical System Backpressure:
Suppressor:
Detectors:
                            Dionex DX500
                            225 |iL
                            9.0mMNa2CO3
                            1.3 mL/min
                            Dionex AG9-HC / AS9-HC, 4 mm
                            2300 psi
                            ASRS-I, external water mode, 100 mA current
                            Suppressed Conductivity Detector, Dionex CD20
                                 Background Conductivity: 24 |j,S
                            Absorbance Detector, Dionex AD20 (10mm cell path)
                                 Set for absorbance at 450 nm (Tungsten lamp)
                            0.7 mL/min
Postcolumn Reactor Coil: knitted, potted for heater, 500 uL internal volume
Postcolumn Heater:            80 °C
Recommended methodtotal analysis time:    25 minutes

Analyte Retention Times and Method Detection Limits  (MDLs):
Postcolumn Reagent Flow:

Analyte
Chlorite(c)
Chlorite(d)
Bromate(c)
Bromate(d)
Bromate(e)
Surrogate: DCA(d)
Bromide(c)
Bromide(d)
Chlorate(c)
Chlorate(d)

Retention Time (b)
(min.)
4.20
4.20
4.85
4.85
5.35
8.50
10.0
10.0
11.0
11.0
MDL
Fortified Cone.
(ng/L)
2.0
2.0
2.0
2.0
0.50

2.0
2.0
2.0
2.0
DETERMINATION
#of
Reps.
8
8
8
8
7

8
8
8
8

MDL
0.45
0.89
0.98
0.71
0.12

0.54
0.69
0.92
0.62
(a) Mention of trade names or commercial products does not constitute endorsement or recommendation
   for use.
(b) Reference to chromatograms in Figure 2 and 3.
(c) Method 317.0 conductivity detection without PCR online.
(d) Method 317.0 conductivity detection with PCR online.
(e) Method 317.0 ONLY bromate bypostcolumn UV/VIS absorbance detection.
                                      317.0-35

-------
TABLE 2.      SINGLE LABORATORY PRECISION IN VARIOUS MATRICES FOR
               BROMATE BY CONDUCTIVITY AND ABSORBANCE DETECTION.
Matrix Detection
Reagent Conductivity
Water
Conductivity
Absorbance
Absorbance
Chlorinated Conductivity
Drinking Water
Conductivity
Absorbance
Absorbance
High Ionic Conductivity
Water
Conductivity
Absorbance
Absorbance
High Conductivity
Organic
Water Conductivity
Absorbance
Absorbance
Fortified
Cone.
(ng/L)
0.50
5.0
0.50
5.0
0.50
5.0
0.50
5.0
0.50
5.0
0.50
5.0
0.50
5.0
0.50
5.0
PRE
# of Reps.
8
8
8
8
8
y(b)
8
8
8
8
8
8
8
8
8
8
CISION
Mean
(ng/L)
< MRL(a)
4.8
0.50
5.4

-------
TABLE 3.   SINGLE-LABORATORY PRECISION AND RECOVERY FOR THE INORGANIC
            DISINFECTION BY-PRODUCTS AND BROMIDE.12
Analyte
Chlorite













Bromate
by
Conductivity











Matrix
RW

HIW

sw

GW

C1W

CDW

O3W

RW

HIW

SW

GW

C1W

CDW

O3W

Unfortified
Cone.
(ng/L)

-------
TABLE 3.   SINGLE-LABORATORY PRECISION AND RECOVERY FOR THE INORGANIC
            DISINFECTION BY-PRODUCTS AND BROMIDE (cont.).12
Analyte Matrix
Bromide RW

HIW

sw

GW

C1W

CDW

O3W

Chlorate RW

HIW

SW

GW

C1W

CDW

O3W

Unfortified
Cone.
Og/L)

-------
TABLE 3.   SINGLE-LABORATORY PRECISION AND RECOVERY FOR THE INORGANIC
            DISINFECTION BY-PRODUCTS AND BROMIDE (cont.).12
Analyte
Surrogate: DCA
(see Note)












Matrix
RW

HIW

sw

GW

C1W

CDW

O3W

Fortified
Cone.
(mg/L)
5.0

5.0

5.0

5.0

5.0

5.0

5.0

#of
Reps.
9

9

9

9

9

9

9

Mean
(mg/L)
5.1
5.0
5.0
5.0
4.9
5.0
5.1
5.1
5.2
5.1
5.0
5.0
5.0
5.1
Mean
%REC
102
99.5
100.
99.2
98.9
99.8
102
103
103
103
100.
101
99.8
101
SD(n-l)
0.93
0.69
0.79
1.76
0.70
1.60
0.50
0.50
1.73
1.12
1.02
1.08
0.70
0.53
%RSD
0.91
0.69
0.79
1.78
0.7
1.61
0.49
0.49
1.68
1.09
1.02
1.07
0.7
0.52
RW =   Reagent Water
HIW =  High Ionic Strength Water
SW =   Surface Water
GW =   Groundwater
C1W =  Chlorinated Drinking Water
CDW = Chlorine Dioxide Treated Drinking Water
O3 W =  Ozonated Drinking Water
NOTE: The surrogate DCA was fortified at 5 mg/L but due to concerns about measuring trace concentrations
       of bromide with such high concentration of the neighboring surrogate peak, the recommended
       fortified concentration for the surrogate has been reduced to 1.00 mg/L.
                                           317.0-39

-------
TABLE 3A.   STABILITY STUDY RESULTS FOR THE INORGANIC DISINFECTION BY-
             PRODUCTS AND BROMIDE.1
                               Unfortified   Fortified
Analyte % Recovery
Analyte
Chlorite





Chlorite






Bromate






Bromate






Preservative Matrix
None RW
HIW
SW
GW
C1W
CDW
O3W
EDA RW
HIW
SW
GW
C1W
CDW
O3W
None RW
HIW
SW
GW
C1W
CDW
O3W
EDA RW
HIW
SW
GW
C1W
CDW
O3W
uonc.

-------
 TABLE 3A. STABILITY STUDY RESULTS FOR THE INORGANIC DISINFECTION
             BY-PRODUCTS AND BROMIDE (cont.).12
                              Unfortified Fortified I
Analyte % Recovery
Analyte
Bromide






Bromide






Chlorate






Chlorate






Preservative Matrix
None RW
HIW
SW
GW
C1W
CDW
O3W
EDA RW
HIW
SW
GW
C1W
CDW
O3W
None RW
HIW
SW
GW
C1W
CDW
O3W
EDA RW
HIW
SW
GW
C1W
CDW
O3W
^UilU.
(|ig/L)

-------
TABLE 4.  INITIAL DEMONSTRATION OF CAPABILITY QC REQUIREMENTS.
  Reference
  Requirement
        Specification and Frequency
Acceptance Criteria
 Sect.  9.2.2
       9.3.1
Initial
Demonstration of
Low System
Background
Analyze a method blank (LRB) and determine
that all target analytes are below % of the
proposed MRL prior to performing the IDC
The LRB
concentration must be
< :/2 of the proposed
MRL
  Sect. 9.2.3
Initial
Demonstration of
Precision (IDP)
Conductivity: analyze 7 replicate LFBs
recommend fortify at 20 ug/L
Absorbance: analyze 7 replicate LFBs
recommend fortify with bromate at 2.0 ug/L
%RSDmustbe<20%
  Sect. 9.2.4
Initial
Demonstration of
Accuracy (IDA)
Calculate average recovery of IDP replicates
Mean % recovery
must be ± 15% of true
value.
  Sect. 9.2.5
Quality Control
Sample (QCS)
Initially and at least quarterly analyze a QCS
from an external/second source
QCS must be ± 20%
of the true value
  Sect. 9.2.6
Method
Detection Limit
(MDL)
Determination
Select a fortifying level at 3-5 times the
estimated instrument detection limit at or lower
than the MRL.  Analyze 7 replicate LFBs
Calculate MDL using equation in Section 9.2.6
- do not subtract blank
  Sect. 9.2.7
Minimum
Reporting Level
(MRL)
MRLs MUST be established for all analytes
during the IDC.
The low CAL
standard can be lower
than the MRL, but the
MRL MUST be no
lower than the low
CAL standard
                                             317.0-42

-------
TABLE 5.  QUALITY CONTROL REQUIREMENTS (SUMMARY).
 Reference
  Requirement
      Specification and Frequency
     Acceptance Criteria
  Sect. 8.3
Sample Holding
Time /
Preservation
Bromate   28 days, refrig. at <6°C /
          EDA Preservation
Bromide   28 days, EDA Permitted
Chlorate   28 days, refrig. at <6°C  / EDA
          Preservation
Chlorite   14 days, refrig. at <6°C  / EDA
          Preservation
Holding time and temperature
must not be exceeded. EDA
added to all samples
    Sect.
   11.1.4.1
  (specific
  to PCR)
Pretreated
Sample
(acidified/Fe [II]
added to remove
chlorite) Holding
Time
ONLY REQUIRED when samples
containing chlorite are pretreated and PCR
is employed to measure trace bromate in
samples.
MAXIMUM PRETREATED SAMPLE
HOLDING TIME: 30 hours
Pretreated sample holding time
must not be exceeded
  Sect. 10.2
Initial
Calibration
Conductivity: generate calibration curve
using at least 5 standards
Absorbance: generate calibration curve
using at least 5 bromate standards
MRL MUST be no lower than
the lowest calibration standard
    Sect.
    10.3.1
Initial
Calibration
Check
Daily, verify calibration of conductivity
detector at the MRL by analyzing an initial
low-level continuing calibration check
standard (ICCS) and a separate low-level
ICCS for the absorbance detector at the
MRL.
Recovery must be 75-125% of
the true value on both detectors
    Sect.
    10.3.2
Continuing
Calibration and
End Calibration
Checks
Alternately analyze separate mid and high
level CCCS/ECCS after every 10 samples
and after the last sample
MRL to 5 x MRL must have 75
125% recovery on both
detectors
For 5 x MRL to highest CCCS
must have 85 - 115% recovery
on both detectors
  Sect. 9.3.1
Laboratory
Reagent Blank
(LRB)
Include LRB with every analysis batch (up
to 20 samples)
Analyze prior to analyzing field samples
All analytes must be
< y2 MRL
    Sect.
   9.3.1.2
  (specific
  to PCR)
PRETREATED
Laboratory
Reagent Blank
(LRB)
REQUIRED in any analysis batch which
includes samples which have been
pretreated to remove chlorite prior to PCR
measurement of trace bromate.
PCR measured bromate
< 1A MRL
  Sect. 9.3.2
Laboratory
Fortified Blank
(LFB)
Laboratory must analyze LFB in each
analysis batch following the ICCS.
Calculate %REC prior to analyzing
samples
LFB recovery if fortified at cone.
from MRL to 5X MRL must be
75- 125%.  For 5X MRL to
highest CCCS must be 85 -
115%. Must have acceptable
recoveries prior to analyzing
samples.  Sample results from
batches that fail LFB are invalid
                                              317.0-43

-------
TABLE 5.  QUALITY CONTROL REQUIREMENTS (SUMMARY CONTINUED).
  Reference
  Requirement
     Specification and Frequency
   Acceptance Criteria
  Sect. 9.3.3
Instrument
Performance
Check (IPC)
Calculate Peak Gaussian Factor (PGF)
using equation (Sect. 9.3.3.1) and
monitor retention time for surrogate in
Initial Calibration Check Standard
(ICCS) each day
PGF must fall between
0.80 and 1.15
Ret. Time (RT) for
surrogate must remain
80% of initial RT when
column was new
  Sect. 9.4.1
Laboratory
Fortified Sample
Matrix (LFM)
 Sect.
 11.1.4.2
Must add known amount of each target
analyte to a minimum of 5% of field
samples or at least one within each
analysis batch for both detectors
LFM must be fortified above the native
level and at no greater than 5 x the
highest field sample concentration
Calculate target analyte recovery using
formula (Sect. 9.4.1.3)
When field samples from chlorine
dioxide plants which contain chlorite
are pretreated prior to the PCR
measurement of trace bromate, an
additional LFM must be prepared for
each pretreated field sample (Sect.
9.4.1.5)
Recovery should be
75 - 125%

If fortified sample fails the
recovery criteria, label
both as suspect/matrix.
  Sect. 9.4.2
Surrogate
Dichloroacetate is added to all blanks,
samples and standards (if measuring by
conductivity and absorbance)
Calculate recovery using formula in
Section 9.4.2
Surrogate recovery must
be 90- 115%.
Samples that fail surrogate
recovery must be
reanalyzed.  If second
analysis fails label result
as suspect/matrix
  Sect. 9.4.3
Field or
Laboratory
Duplicates
Analyze either a field or laboratory
duplicate for a minimum of 5% of field
samples or at least one within each
analysis batch for both detectors
Calculate the relative percent difference
(RPD) using formula in Section9.4.3.1
The RPD for
concentrations at MRL to
5 x MRL should be ± 20%
on both detectors, and ±
10% on both detectors
for concentrations at 5 x
MRLtohighestCCCS. If
this range is exceeded,
label both as
suspect/matrix
                                            317.0-44

-------
TABLE 6.  EXAMPLE SAMPLE ANALYSIS BATCH WITH QUALITY CONTROL
         REQUIREMENTS
Injection
#
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
Sample
Description
Laboratory reagent blank (LRB)
ICCS conductivity detector (5.0 ug/L)
ICCS absorbance detector (0.5 ngfL)
Laboratory Fortified Blank (LFB) -
conductivity detector
LFB - absorbance detector
Field sample 1
Field sample 1 - Laboratory Duplicate (LD) (a)
Field sample 2
Field sample 2 - Laboratory Fortified Matrix (LFM) (a) at
concentrations specific for conductivity detector
Field sample 2 - LFM specific for trace bromate on the
absorbance detector
Field sample 3
Field sample 4
Field sample 5
Field sample 6
Field sample 7
Field sample 8
Field sample 9
Field sample 1 0
CCCS conductivity detector (75.0 |j,g/L)
CCCS absorbance detector (5.0 (ig/L)
Field sample 1 1
Field sample 12
Acceptance
Criteria
< 1A MRL
3.75to6.25^g/L
0.375 to 0.625 ng/L
± 25 % fortified level
± 25 % fortified level

± 15%RPD

± 25% fortified level
± 25% fortified level








63. 8 to 86.3 ug/L
4.25 to 5.75 ug/L


                                 317.0-45

-------
23
24
25
26
27
28
29
30
31
32
33
34
Field sample 1 3
Field sample 14 - (finished water from PWS using chlorine
dioxide)
PretreatedLRB (Section 9.3.1.2) using the acid/Fe(II)
chlorite removal procedure (Section 1 1.1.4)
Field sample 14 (b) - (finished water from PWS using
chlorine dioxide) pretreated with acid/Fe(II) (Section
11.1.4)
Field sample 14 - (finished water from PWS using chlorine
dioxide) LFM specific for trace bromate on the absorbance
detector, pretreated with acid/Fe(II) (Section 1 1.1.4.2)
Field sample 1 5
Field sample 1 6
Field sample 1 7
Field sample 1 8
Field sample 19(b)
ECCS conductivity detector (500.0 (ig/L)
ECCS absorbance detector (15.0 ng/L)


< y2 MRL

± 25% fortified level





425 to 575 ng/L
12. 8 to 17.3 (ig/L
(a)   If no analytes are observed above the MRL for a sample, an alternate sample which contains reportable
    values should be selected as the laboratory duplicate. Alternately, the LFM can be selected and
    reanalyzed as the laboratory duplicate ensuring the collection of QC data for precision.

(b)   Field sample #19 was the final field sample permitted in this batch but 20 total field samples were
    analyzed.
    Field sample #14 was analyzed both initially and as a acid/Fe (II) pretreated sample, therefore, it
    accounted for two "field sample analyses" toward the maximum of twenty in an analysis batch (Section
    3.1).
                                            317.0-46

-------
      System Configuration for EPA Method 317.0
     Eluent
 Auto sampler
    DX -500
                     Injection loop
AG9-HC
AS9-HC
                    PC10PCR
                     Reservoir
Electrolytic
Suppressor
                                       Abs or bance
                                         Detector
                                 Conductivity
                                   Detector
                                                   Knitted Reaction Coil
                                                   PCH-2 Heater (ffi 60 °C
                            Effluent to waste
Figure 1:   Schematic detailing the configuration of postcolumn hardware addition to an ion
         chromatograph. Mention of trade names or commercial products does not constitute
         endorsement or recommendation for use.  If the requirements found in Section 9 are met,
         equivalent products or hardware can be employed.

         NOTE: In a typical Method 300.1 hardware configuration, a backpressure coil is included after
         the conductivity cell as part of the waste stream when this manufacturer's equipment is used.
         These backpressure coils are not required when the Method 317.0 instrument configuration is
         employed since the additional PCR system components, placed in-line, function in the same
         capacity and provide sufficient backpressure.
                                   317.0-47

-------
   0.500r
   0.400 --
   0.100 -
  -0.100 --
                                  Surrogate:
                                  DCA  X
                        bromate
                    chlorite
                                             bromide  chlorate
                                                                  Conductivity Detector
                2.50        5.00        7.50       10.00       12.50       15.00       17.50       20.00       22.50       25.00
                                                Minutes
2.00x10-V
1.50X10"3--
1.00x10"*--
5.00x10"*--
bromate
   I
                                                                  UV7VIS Detector at 450 nm
                 2.50        5.00        7.5D
                                                10.00
                                                Minutes
                                                           12.50       15.00       17,50      20.00       22.50       25.00
        Figure 2:   Reagent water fortified with inorganic disinfection by-products and bromide at 10
                     ug/L.
                                                  317.0-48

-------
   0.400
   0,200
   0.100--
   -0.200 L
        0
                                    Surrogate^
                                    DCA
                         bromate
                 2.50        S.OO
                                                                    Conductivity Detector
                                       7.50        10.00
                                                 Minutes
                                                             12.SO       15.00       17.50       20.00       22.60        25.00
 1.00x10 --
                            bromate
                                                                    UV/VIS Detector at 450 nm
-8.00x10 -I—f
       0
                 2.50        S.OO        7.50
                                                  10.00       12.60
                                                 Minutes
                                                                        19MU       17.50       20.00       22.50       25.00
         Figure 3:   Chlorinated tap water fortified with bromate at 2.0 ug/L.
                                                   317.0-49

-------