EPA Document #: 815-B-07-002
METHOD 536 DETERMINATION OF TRIAZINE PESTICIDES AND THEIR
DEGRADATES IN DRINKING WATER BY LIQUID
CHROMATOGRAPHY ELECTROSPRAY IONIZATION TANDEM
MASS SPECTROMETRY (LC/ESI-MS/MS)
Version 1.0
October 2007
G.A. Smith and B.V. Pepich (Shaw Environmental, Inc.)
D. J. Munch (U.S. EPA, Office of Ground Water and Drinking Water)
TECHNICAL SUPPORT CENTER
OFFICE OF GROUND WATER AND DRINKING WATER
U. S. ENVIRONMENTAL PROTECTION AGENCY
CINCINNATI, OHIO 45268
536-1
-------
METHOD 536
DETERMINATION OF TRIAZINE PESTICIDES AND THEIR DEGRADATES IN DRINKING
WATER BY LIQUID CHROMATOGRAPHY ELECTROSPRAY IONIZATION TANDEM
MASS SPECTROMETRY (LC/ESI-MS/MS)
1. SCOPE AND APPLICATION
1.1 This is a liquid chromatography electrospray ionization tandem mass spectrometry (LC/ESI-
MS/MS) method for the determination of triazine pesticides and their degradation products in
finished drinking waters. This method requires the use of MS/MS in Multiple Reaction
Monitoring (MRM) mode to enhance selectivity. Precision and accuracy data have been
generated for the detection of the triazine pesticides and their degradates in reagent water and
finished drinking water from both groundwater and surface water sources. The single
laboratory Lowest Concentration Minimum Reporting Level (LCMRL) has also been
determined in reagent water.1 The following compounds can be determined using this
method:
Chemical Abstract Services
Analvte Registry Number (CASRN)
Atrazine 1912-24-9
Atrazine-desethyl 6190-65-4
Atrazine-desethyl-desisopropyl 3397-62-4
Atrazine-desisopropyl 1007-28-9
Cyanazine 21725-46-2
Propazine 139-40-2
Simazine 122-34-9
1.2 The MRM mass spectrometry conditions described in this method were developed using a
conventional LC/MS/MS system.
1.3 The Minimum Reporting Level (MRL) is the lowest analyte concentration that meets the Data
Quality Objectives (DQOs), which are based on the intended use of this method. The single
laboratory LCMRL is the lowest true concentration for which the future recovery is predicted
to fall between 50-150% recovery with 99% confidence. Single laboratory LCMRLs for the
analytes in this method ranged from 0.020 - 0.14 |J,g/L, and are listed in Table 4. The
procedure used to determine the LCMRL is described elsewhere.1
1.4 Laboratories using this method are not required to determine an LCMRL, but they will need
to demonstrate that their laboratory Minimum Reporting Level (MRL) for this method meets
the requirements described in Section 9.2.4.
1.5 Detection limit (DL) is defined as the statistically calculated minimum concentration that can
be measured with 99% confidence that the reported value is greater than zero.2 The DL is
dependent on sample matrix, fortification concentration, and instrument performance.
536-2
-------
Determining the DL for analytes in this method is optional (Sect. 9.2.6). DLs for method
analytes fortified into reagent water ranged from 0.010 - 0.043 ng/L. These values are also
provided in Table 4.
1.6 This method is intended for use by analysts skilled in the operation of LC/MS/MS
instrumentation and the interpretation of the associated data.
1.7 METHOD FLEXIBILITY - In recognition of technological advances in analytical systems
and techniques, the laboratory is permitted to modify the separation technique, LC column,
mobile phase composition, LC conditions and MS conditions (Sect. 6.10, 9.4, 10.2.2, 10.2.3,
10.2.4, and 12.1). Changes may not be made to sample collection and preservation (Sect.
8) or to the quality control requirements (Sect. 9). Method modifications should be
considered only to improve method performance. Modifications that are introduced in the
interest of reducing cost or sample processing time, but result in poorer method performance,
may not be used. The analytes must be resolved chromatographically in order to avoid
the effects of matrix suppression or enhancement and to provide analyte specificity. In
all cases where method modifications are proposed, the analyst must perform the procedures
outlined in the Initial Demonstration of Capability (IDC, Sect. 9.2), verify that all QC
acceptance criteria in this method (Tables 9 and 10) are met, and verify method performance
on a real sample matrix (Sect. 9.4).
2. SUMMARY OF METHOD
2.1 Samples are pH adjusted and dechlorinated with ammonium acetate and protected from
microbial degradation using sodium omadine during sample collection. Prior to analysis,
isotopically enriched analytes [Atrazine-d5(ethyl-d5), Atrazine-desethyl-d7, Atrazine-
desethyl-desisopropyl(13C3), Atrazine-desisopropyl-d5(ethyl-d5), Cyanazine-d5(N-ethyl-d5),
Propazine-dl4(di-isopropyl-dl4), and Simazine-dlO(diethyl-dlO)] are added to the samples
as internal standards. An aliquot of the sample is injected without cleanup or concentration
onto a chromatographic column (CIS). The analytes are separated using an ammonium
acetate-buffered binary mobile phase and subsequently detected by positive electrospray
ionization tandem mass spectrometry. The concentration of each identified component is
determined using the internal standard technique.
3. DEFINITIONS
3.1 ANALYSIS BATCH - A sequence of samples, which are analyzed within a 30-hour period
and include no more than 20 field samples. Each Analysis Batch must also include all
required QC samples, which do not contribute to the maximum field sample total of 20. The
required QC samples include:
Laboratory Reagent Blank (LRB),
Continuing Calibration Check Standards (CCCs),
Laboratory Fortified Sample Matrix (LFSM), and
Laboratory Fortified Sample Matrix Duplicate or Laboratory Duplicate (LFSMD or LD).
3.2 CALIBRATION STANDARD (CAL) - A solution of the target analytes prepared from the
primary dilution standard solution(s) or stock standard solution(s) and the internal standards.
The CAL solutions are used to calibrate the instrument response with respect to analyte
concentration.
536-3
-------
3.3 CONTINUING CALIBRATION CHECK (CCC) - A calibration standard containing the
method analytes and internal standards which is analyzed periodically to verify the accuracy
of the existing calibration.
3.4 DETECTION LIMIT (DL) - The minimum concentration of an analyte that can be
identified, measured and reported with 99% confidence that the analyte concentration is
greater than zero. This is a statistical determination (Sect. 9.2.6), and accurate quantitation is
not expected at this level.2
3.5 INTERNAL STANDARD (IS) - A pure compound added to all standard solutions and field
samples in a known amount. It is used to measure the relative response of the method
analyte. The internal standard must be a compound that is not a sample component.
3.6 ION SUPRESSIGN/ENHANCEMENT - An observable loss or increase in target analyte
response in complex (field) samples as compared to the response obtained in standard
solutions.
3.7 LABORATORY DUPLICATES (LDs) - Two sample aliquots (LDi and LD2), taken in the
laboratory from a single sample bottle, and analyzed separately with identical procedures.
Analyses of LDi and LD2 indicate precision associated specifically with the laboratory
procedures by removing variation contributed from sample collection, preservation and
storage procedures.
3.8 LABORATORY FORTIFIED BLANK (LFB) - An aliquot of reagent water or other blank
matrix to which a known quantity of the method analytes is added. The LFB is analyzed
exactly like a sample including the preservation procedures in Section 8. Its purpose is to
determine whether the methodology is in control, and whether the laboratory is capable of
making accurate measurements.
3.9 LABORATORY FORTIFIED SAMPLE MATRIX (LFSM) - An aliquot of a field sample to
which a known quantity of the method analytes is added. The LFSM is processed and
analyzed exactly like a sample, and its purpose is to determine whether the sample matrix
contributes bias to the analytical results. The background concentration of the analytes in the
sample matrix must be determined in a separate aliquot and the measured value in the LFSM
corrected for background concentrations.
3.10 LABORATORY FORTIFIED SAMPLE MATRIX DUPLICATE (LFSMD) - A second
aliquot of the field sample used to prepare the LFSM which is fortified and analyzed
identically to the LFSM. The LFSMD is used instead of the Laboratory Duplicate to assess
method precision and accuracy when the occurrence of the target analytes is infrequent.
3.11 LABORATORY REAGENT BLANK (LRB) - An aliquot of reagent water or other blank
matrix that is treated exactly as a sample including exposure to all storage containers, buffers,
preservatives and internal standards. The LRB is used to determine if the method analytes or
other interferences are present in the laboratory environment, the reagents, or the apparatus.
536-4
-------
3.12 LOWEST CONCENTRATION MINIMUM REPORTING LEVEL (LCMRL) - The single
laboratory LCMRL is the lowest true concentration for which the future recovery is predicted
to fall between 50-150% recovery with 99% confidence.1
3.13 MATERIAL SAFETY DATA SHEETS (MSDS) - These sheets contain written information
provided by vendors concerning a chemical's toxicity, health hazards, physical properties,
fire, and reactivity data including storage, spill, and handling precautions.
3.14 MINIMUM REPORTING LEVEL (MRL) - The minimum concentration that can be
reported by a laboratory as a quantified value for the target analyte in a sample following
analysis. This concentration must meet the criteria defined in Section 9.2.4 and must be no
lower than the concentration of the lowest calibration standard for the target analyte.
3.15 MULTIPLE REACTION MONITORING (MRM) - A mass spectrometric technique in
which a precursor ion is first isolated then subsequently fragmented into product ion(s).
Quantitation is accomplished by monitoring a specific product ion. As described in Section
10.2.2, MS parameters must be optimized for each precursor ion (Sect. 3.16) and product ion
(Sect. 3.18).
3.16 PRECURSOR ION - For the purpose of this method, the precursor ion is the protonated
molecule ([M+H]+) of the target analyte ([M]). In MS/MS, the precursor ion is mass selected
and fragmented by collisionally activated dissociation to produce distinctive product ions of
smaller m/z.
3.17 PRIMARY DILUTION STANDARD SOLUTION (PDS) - A solution containing the
method analytes prepared in the laboratory from stock standard solutions and diluted as
needed to prepare calibration standards and other analyte solutions.
3.18 PRODUCT ION - For the purpose of this method, a product ion is one of the fragment ions
produced in MS/MS by collisionally activated dissociation of the precursor ion.
3.19 QUALITY CONTROL SAMPLE (QCS) - A solution containing the method analytes at a
known concentration which is obtained from a source external to the laboratory and different
from the source of calibration standards. It is used to check laboratory performance with
externally prepared test materials.
3.20 REAGENT WATER (RW) - Purified water which does not contain any measurable quantity
of the target analytes or interfering compounds at or above 1/3 the MRL for the target
analytes.
3.21 STOCK STANDARD SOLUTION (SSS) - A concentrated solution containing one or more
of the method analytes that is prepared in the laboratory using assayed reference materials or
purchased from a reputable commercial source, so that the concentration and purity of
analytes are traceable to certificates of analysis.
4. INTERFERENCES
4.1 All glassware must be meticulously cleaned. Wash glassware thoroughly and rinse with
reagent water.
536-5
-------
4.2 Method interferences may be caused by contaminants in solvents, reagents (including reagent
water), sample bottles and caps, and other sample processing hardware. These interferences
may lead to discrete artifacts and/or elevated baselines in the chromatograms. All laboratory
reagents and equipment must be routinely demonstrated to be free from interferences (less
than 1/3 the MRL for the target analytes) under the conditions of the analysis. This may be
accomplished by analyzing LRBs as described in Section 3.11. Subtracting blank values
from sample results is not permitted.
4.3 Equipment used for sample collection and storage has the potential to introduce
interferences. The potential for interferences from these devices must be investigated during
the Initial Demonstration of Capability (Sect. 9.2) by preparing and analyzing a Laboratory
Reagent Blank (LRB). This procedure should be repeated each time a new brand or lot of
equipment is used to ensure that background contamination does not interfere with the
identification and quantitation of the analytes.
4.4 Matrix interferences may be caused by contaminants that are present in the sample. The
extent of matrix interferences will vary considerably from source to source depending upon
the nature of the water.
4.4.1 ION SUPPRESSION/ENHANCEMENT - Field samples containing moderate-to-high
levels of organic carbon or dissolved solids may cause suppression or enhancement of
instrumental response for one or more of the analytes. Suppression and enhancement
effects are introduced during the ionization process in the electrospray source when a co-
eluting contaminant influences the ionization of the analyte of interest. MS/MS methods
are not immune to suppression or enhancement effects because the fragmentation process
occurs after ionization. Isotopically enriched internal standards can help mitigate the
effect of matrix-induced suppression or enhancement, but good performance is best
achieved by optimizing the chromatographic separation.
4.4.2 INTERFERING COMMON IONS - While MS/MS detection offers increased selectivity
relative to MS mode detection, interference can still occur in complex samples if
common ions are present in co-eluting components. Due to a lack of LC/MS/MS
reference spectral libraries, optimized chromatographic separation is required for mass
resolution of the analytes of interest in the method.
5. SAFETY
5.1 The toxicity or carcinogenicity of each reagent used in this method has not been precisely
defined. Each chemical should be treated as a potential health hazard, and exposure to these
chemicals should be minimized. Each laboratory is responsible for maintaining an awareness
of OSHA regulations regarding safe handling of chemicals used in this method. A reference
file of MSDSs should be made available to all personnel involved in the chemical analysis.
5.2 Pure standard materials and stock standard solutions of the method compounds should be
handled with suitable protection for skin, eyes, etc.3"5
536-6
-------
6. EQUIPMENT AND SUPPLIES (References to specific brands or catalog numbers are included
as examples only and do not imply endorsement of the product. This does not preclude the use of
other vendors or suppliers.)
6.1 SAMPLE CONTAINERS - Pre-cleaned, amber vials (40 mL or larger) fitted with
PTFE/silicone septa and polypropylene screw caps (Fisher Cat. No. 02-912-377 or
equivalent).
6.2 VIALS - Amber 2-mL glass autosampler vials with PTFE/silicone septa for storing standards
(Fisher Cat. No. 03-375-20B or equivalent).
6.3 MICRO SYRINGES - Suggested sizes include 10, 25, 50, 100, 250, and 500 |iL.
6.4 VOLUMETRIC FLASKS - Class A, suggested sizes include 5, 10, 50, 100, 250, and 500
mL for preparation of reagents and standards.
6.5 VOLUMETRIC PIPETTES - Class A, suggested sizes include 1, 4, 5, 10, 20, and 25 mL.
6.6 ANALYTICAL BALANCE - Capable of weighing to the nearest 0.0001 g.
6.7 DESOLVATION GAS - High purity compressed gas (e.g., nitrogen or zero-air) used for
desolvation in the mass spectrometer. The specific type of gas, purity and pressure
requirements will depend on the instrument manufacturers' specifications.
6.8 COLLISION GAS - High purity compressed gas (e.g., nitrogen or argon) used in the
collision cell of the mass spectrometer. The specific type of gas, purity and pressure
requirements will depend on the instrument manufacturer's specifications.
6.9 DISPOSABLE PASTEUR PIPETS - 9" borosilicate glass, used to transfer samples to
autosampler vials (Fisher Cat. No. 13-678-20C or equivalent).
6.10 LIQUID CHROMATOGRAPHY ELECTROSPRAY TANDEM MASS SPECTROMETRY
SYSTEM (LC/ESI-MS/MS)
6.10.1 LC COLUMN-LC column, Waters Xterra® RP18, 2.1 x 150 mm, 3.5 |im dp (Waters
Part No. 186000410 or equivalent). Any column that provides adequate resolution, peak
shape, capacity (Sect. 10.2.3 and 10.2.4), accuracy and precision (Sect. 9), and does not
result in suppression or enhancement of analyte response (Sect. 13.4.2) may be used.
6.10.2 LC SYSTEM -The LC system (Waters 2695 or equivalent) must provide consistent
sample injection volumes and be capable of performing binary linear gradients at a
constant flow rate.
6.10.3 MASS SPECTROMETER -The MS (Micromass QuattroMicro or equivalent) must be
capable of providing electrospray ionization with positive ion detection. The system
must be capable of performing MS/MS to produce unique product ions (Sect. 3.18) for
the target analytes within specified retention time windows. Data presented in Section
17, Tables 4-8 were collected using an ESI-MS/MS system.
536-7
-------
6. 10.4 DATA SYSTEM - An interfaced data system is required to acquire, store, and output MS
data. The computer software should have the capability of processing stored LC/MS/MS
data by recognizing a chromatographic peak within a given retention time window. The
software must allow integration of the ion abundance of any specific ion between
specified times or scan number limits. The software must be able to construct a linear
regression or quadratic calibration curve, and calculate analyte concentrations using the
internal standard technique.
7. REAGENTS AND STANDARDS
7. 1 REAGENTS AND SOLVENTS - Reagent grade or better chemicals should be used. Unless
otherwise indicated, it is intended that all reagents will conform to the specifications of the
Committee on Analytical Reagents of the American Chemical Society (ACS), where such
specifications are available. Other grades may be used, as long as the reagent is of
sufficiently high purity to permit its use without lessening the quality of the determination.
7.1.1 METHANOL - (CH3OH, CAS# 67-56-1) - High purity, demonstrated to be free of
analytes and interferences (Fisher GC Resolv grade or equivalent).
7.1.2 AMMONIUM ACETATE - (NFLAF^, CAS# 63 1 -6 1 -8) - High purity, demonstrated
to be free of analytes and interferences (Fisher Brand HPLC grade or equivalent).
7.1.3 5 mM AMMONIUM ACETATE MOBILE PHASE - To prepare 1 L of mobile phase,
add 0.3854 g ammonium acetate to 1 L of reagent water (final pH ~ 6.5). This solution is
prone to volatility losses and should be replaced every 48 hours.
7. 1 .4 2.0 M AMMONIUM ACETATE/REAGENT WATER CONCENTRATED STOCK -
Add 38.5 g of ammonium acetate to a 250-mL volumetric flask. Dilute to volume with
reagent water. An aliquot of this concentrated stock solution is diluted 1 : 100 in all
calibration standards and samples to yield a final concentration of 20 mM (1.5 g/L).
7.1.5 SODIUM OMADINE® (Sodium 2-pyridinethiol-l -oxide, CAS# 3811-73-2) -
(C5H4NOSNa) - 40 wt.% in water (Omadine is a registered trademark of Arch
Chemicals, Inc.).
7.1.6 SODIUM OMADINE CONCENTRATED STOCK SOLUTION (32 g/L) - Add 20 mL
of the 40 wt.% sodium omadine in water to a 250-mL volumetric flask. Dilute to volume
with reagent water. The concentrated stock solution is diluted 1 :500 in all calibration
standards and samples to yield a final concentration of 64 mg/L.
7.2 STANDARD SOLUTIONS - When a compound's purity is assayed to be 96 percent or
greater, the weight can be used without correction to calculate the concentration of the stock
standard. Solution concentrations listed in this section were used to develop this method
and are included only as an example. Stock standard solutions are estimated to be stable
for six months. Any fortified or dilute solutions made from the stock standards are stable for
at least 60 days provided they are stored at a temperature < 6 °C and the stock standard
solutions have not exceeded their six month stability period. Although estimated stability
times for standard solutions are given, laboratories should use standard QC practices to
determine when their standards need to be replaced.
536-8
-------
7.2.1 INTERNAL STANDARD SOLUTIONS - This method uses a separate isotopically
enriched internal standard for each target analyte, specifically Atrazine-d5(ethyl-d5)
(CAS# 163165-75-1), Atrazine-desethyl-d7, Atrazine-desethyl-desisopropyl(13C3),
Atrazine-desisopropyl-d5(ethyl-d5), Cyanazine-d5(N-ethyl-d5), Propazine-dl4(di-
isopropyl-d!4), and Simazine-dlO(diethyl-dlO) (CAS# 220621-39-6). It should be noted
that all of the internal standards except atrazine-desethyl-desisopropyl (13C3) are
commercially available; the atrazine-desethyl-desisopropyl (13C3) internal standard used
during method development was custom synthesized by Cambridge Laboratories.
7.2.1.1 INTERNAL STANDARD STOCK STANDARDS (ISSS) (500 ng/mL, except as
noted) - Prepare the stock standards individually by weighing 2.5 mg of the solid
standards into tared 5-mL volumetric flasks and diluting to volume with methanol.
Atrazine-desethyl-desisopropyl (13C3) and Simazine-dlO (diethyl-dlO) exhibit
limited solubility in methanol and must be prepared as < 100 |j,g/mL stock solutions.
To prepare the 100 |j,g/mL stock solutions, weigh 2.5 mg of the solid standards into
tared 25-mL volumetric flasks and dilute to volume with methanol. Sonication may
be required to achieve full dissolution of the standards.
7.2.1.2 INTERNAL STANDARD PRIMARY DILUTION STANDARD (PDS) (1.0
Hg/mL) - Prepare the internal standard PDS by adding enough of each ISSS to a
volumetric flask partially filled with methanol to make the final concentrations near
1.0 |j,g/mL when filled to the mark with methanol. During method development,
addition of 5 |jL of the internal standard PDS to each 1.0-mL sample produced a
final concentration of 5 |ig/L. Analysts are permitted to use other PDS
concentrations and volumes provided all field samples and CAL standards contain
the same amount of internal standard, the concentration of the internal standard
added provides adequate signal to maintain precision, and the volume added has a
negligible effect on the final concentration.
7.2.2 ANALYTE STANDARD SOLUTIONS - Obtain the analytes listed in the table in
Section 1.1 as neat standards. Prepare the Analyte Stock and Primary Dilution Standards
as described below.
7.2.2.1 ANALYTE STOCK STANDARD SOLUTION (500 ng/mL, except as noted) -
Prepare the stock standards individually by weighing 2.5 mg of the solid standards
into tared 5-mL volumetric flasks and diluting to volume with methanol. Atrazine-
desethyl-desisopropyl and Simazine exhibit limited solubility in methanol and must
be prepared as < 100 |j,g/mL stock solutions. To prepare the 100 |j,g/mL stock
solutions, weigh 2.5 mg of the solid standards into tared 25-mL volumetric flasks
and dilute to volume with methanol. Sonication may be required to achieve full
dissolution of the standards.
7.2.2.2 ANALYTE PRIMARY DILUTION SOLUTION (PDS) (2.5 - 5 ng/mL) - The
analyte PDS is prepared by dilution of the Analyte Stock Standard solutions and is
used to prepare the calibration standards and to fortify the LFBs, LFSMs and
LFSMDs with the target analytes.
536-9
-------
CALIBRATION (CAL) STANDARDS - Prepare a procedural calibration curve from
dilutions of the analyte PDS in reagent water. Transfer the standard to a proper storage
container and add appropriate amounts of the ammonium acetate and sodium omadine
concentrated stocks (Sect. 7.1.4 and 7.1.6) as shown in the table below. The target
analyte concentrations found in Tables 5-7 can be used as a starting point for determining
the calibration range. An example of the dilutions used to prepare the CALs that were
used to collect data in Section 17 is shown below. The lowest concentration calibration
standard must be at or below the MRL, which may depend on system sensitivity. The
CAL standards may also be used as CCCs. If stored, the aqueous standards must be
refrigerated in the same manner as the samples. A constant amount of the internal
standard is added to each prepared Calibration Standard This is accomplished for each
standard by taking 1.0 mL of the final calibration standard containing the 20 mM
ammonium acetate and 64 mg/L sodium omadine, and placing it in a 2.0-mL autosampler
vial and adding the appropriate volume of Internal Standard PDS (Sect. 7.2.1.2).
CAL
Level
1
2
3
4
5
6
Analyte
PDS
Cone.
(|ig/mL)
5
5
5
5
5
5
Analyte
PDS
Volume
(HL)
5
10
20
20
25
10
Ammonium
Acetate Stock
Volume
(mL)
1.0
1.0
1.0
0.5
0.5
0.1
Sodium
Omadine Stock
Volume *
(mL)
0.2
0.2
0.2
0.1
0.1
0.02
Final CAL
Std. Volume
(mL)
100
100
100
50
50
10
Final CAL
Std. Cone.
(Hg/L)
0.25
0.50
1.00
2.00
2.50
5.00
Ammonium acetate stock (2.0 M) and sodium omadine stock (32 g/L) are added after the
standards have been diluted to volume and transferred to storage containers.
7.2.3.1 This is a procedural calibration technique. If an alternate set of dilutions are used to
prepare the calibration standards, the analyst must ensure that the dilution factor
used to prepare the calibration standards remains consistent with the dilution factor
during sample collection (Sect. 8.1.1). The dilution factor is currently 101.2/100 in
the table above and in Section 8.1. The analyst must also ensure that all CAL
standards, QC and field samples contain the preservation reagents at their proper
concentrations (Sect. 8.1.1).
8. SAMPLE COLLECTION, PRESERVATION, AND STORAGE
8.1 SAMPLE COLLECTION
8.1.1. Prior to shipment to the field, ammonium acetate and sodium omadine must be added to
each amber bottle fitted with a PTFE-lined screw cap. If using the suggested 40-mL vials
to collect a 40-mL aliquot, add 400 |jL of the ammonium acetate concentrated stock
(Sect. 7.1.4) and 80 |jL of the concentrated sodium omadine stock (Sect. 7.1.6). If other
collection volumes are used, adjust the amount of preservation reagent so that the final
concentrations of ammonium acetate and sodium omadine in the sample containers are
1.5 g/L and 64 mg/L, respectively. Cap the vials to avoid evaporation of the preservation
reagents.
536 -10
-------
8.1.2 When sampling from a cold water tap, open the tap and allow the system to flush until the
water temperature has stabilized (usually approximately 3 to 5 minutes). Collect a
representative sample from the flowing system using a beaker of appropriate size. Use
this bulk sample to generate individual samples as needed. Transfer a volume of at least
40 mLs into each collection container, cap the container, and invert it several times to
mix the sample with the preservation reagents. Care should be taken not to overfill the
vial and flush out the preservation reagents. Samples do not need to be headspace free.
8.1.3 When sampling from an open body of water, fill a beaker with water sampled from a
representative area. Use this bulk sample to generate individual samples as needed.
8.2 SAMPLE SHIPMENT AND STORAGE - Samples must be chilled during shipment and
must not exceed 10 °C during the first 48 hours after collection. Samples should be
confirmed to be at or below 10 °C when they are received at the laboratory. Samples stored
in the lab must be stored at or below 6 °C until analysis. Samples should not be frozen.
8.3 SAMPLE HOLDING TIMES - Samples should be analyzed as soon as possible. Samples
that are collected and stored as described in Sections 8.1 and 8.2 may be stored prior to
analysis for a maximum of 28 days.
9. QUALITY CONTROL
9.1 Quality Control (QC) requirements include the Initial Demonstration of Capability (IDC) and
ongoing QC requirements that must be met when preparing and analyzing field samples.
This section describes each QC parameter, its required frequency, and the performance
criteria that must be met in order to meet EPA quality objectives. The QC criteria discussed
in the following sections are summarized in Section 17, Tables 9 and 10. These QC
requirements are considered the minimum acceptable QC criteria. Laboratories are
encouraged to institute additional QC practices to meet their specific needs.
9.2 INITIAL DEMONSTRATION OF CAPABILITY (IDC) - The IDC must be successfully
performed prior to analyzing any field samples. Prior to conducting the IDC, the analyst
must meet the calibration requirements outlined in Sections 10.2 and 10.3. Requirements for
the initial demonstration of laboratory capability are described in the following sections and
are summarized in Table 9.
9.2.1 DEMONSTRATION OF LOW SYSTEM BACKGROUND - Analyze a LRB and
confirm that the blank is reasonably free of contamination and that the criteria in Section
9.3.1 are met.
NOTE: It is good laboratory practice to include a blank in the calibration of any
instrument. The method should also be checked for carry-over by analyzing a RW blank
immediately following the highest CAL standard. If this RW sample does not meet the
criteria outline in Section 9.3.1, then carry-over is present and should be identified and
eliminated.
9.2.2 DEMONSTRATION OF PRECISION - Prepare and analyze 7 replicate LFBs. These
samples should be fortified near the midrange of the initial calibration curve.
536-11
-------
Ammonium acetate and sodium omadine must be added to the samples as described in
Section 8.1. The percent relative standard deviation (%RSD) of the concentrations of the
replicate analyses must be < 20% for all target analytes.
n / ™ o^ Standard Deviation of Measured Concentrations , ^
%RSD = xlOO
Average Concentration
9.2.3 DEMONSTRATION OF ACCURACY - Using the same sets of replicate data generated
for Section 9.2.2, calculate the average percent recovery. The average percent recovery
of the replicate analyses must be within + 30% of the true value.
„._, Average Measured Concentration ,^
% Recovery = x 100
Fortified Concentration
9.2.4 MINIMUM REPORTING LEVEL (MRL) CONFIRMATION - Establish a target
concentration for the MRL based on the intended use of the method. The lowest
calibration standard used to establish the initial calibration (as well as the low-level
Continuing Calibration Check) must be at or below the concentration of the MRL.
Establishing the MRL concentration too low may cause repeated failure of ongoing QC
requirements. Confirm or validate the MRL following the procedure outlined below.
9.2.4.1 Fortify and analyze seven replicate LFBs at or below the proposed MRL
concentration. All samples must be fortified and processed using the sample
collection protocols described in Section 8.1. Calculate the mean (Mean) and
standard deviation (S) for these replicates. Determine the Half Range for the
Prediction Interval of Results (HRPIR) using the equation below
HRPm = 3.9638
where S is the standard deviation, and 3.963 is a constant value for seven replicates.1
9.2.4.2 Confirm that the upper and lower limits for the Prediction Interval of Results (PIR =
Mean +_ HRpiR) meet the upper and lower recovery limits as shown below.
The Upper PIR Limit must be < 150 percent recovery.
Mean + HRPIP
— x 100 < 150%
Fortified Concentration
The Lower PIR Limit must be > 50 percent recovery.
Mean - HRPIP
— —— — x 100> 50%
r ortijied L oncentration
536 -12
-------
9.2.4.3 The MRL is validated if both the Upper and Lower PIR Limits meet the criteria
described above. If these criteria are not met, the MRL has been set too low and
must be determined again at a higher concentration.
9.2.5 QUALITY CONTROL SAMPLE (QCS) - Analyze a mid-level Quality Control Sample
(Sect. 9.3.7) to confirm the accuracy of the calibration curve fit.
9.2.6 DETECTION LIMIT DETERMINATION (optional) - While DL determination is not a
specific requirement of this method, it may be required by various regulatory bodies
associated with compliance monitoring. It is the responsibility of the laboratory to
determine ifDL determination is required based upon the intended use of the data.
Analyses for this procedure should be done over at least 3 days. Prepare at least 7
replicate LFBs. Ammonium acetate and sodium omadine must be added to the samples
as described in Section 8.1. Use the solutions described in Section 7.2 to fortify at a
concentration estimated to be near the DL. This concentration may be estimated by
selecting a concentration at 2-5 times the noise level. Analyze the seven replicates
through all steps of Section 11.
NOTE: If an MRL confirmation data set meets these requirements, a DL may be
calculated from the MRL confirmation data, and no additional analyses are necessary.
Calculate the DL using the following equation:
DL = St(n-l,l-a = 0.99)
where
t(n-i,i-a = 0.99)= Student's t value for the 99% confidence level with n-1 degrees of
freedom (for seven replicate determinations, the Student's t value
is 3.143 at a 99% confidence level),
n = number of replicates, and
S = standard deviation of replicate analyses.
NOTE: Do not subtract blank values when performing DL calculations.
9.3 ONGOING QC REQUIREMENTS - This section describes the ongoing QC criteria that
must be followed when processing and analyzing field samples. Table 10 summarizes these
requirements.
9.3.1 LABORATORY REAGENT BLANK (LRB) - A LRB is analyzed during the IDC and is
required with each Analysis Batch (Sect. 3.1) to confirm that background contaminants
are not interfering with the identification or quantitation of the target analytes. If within
the retention time window of any analyte, the LRB produces a peak that would prevent
the determination of that analyte, determine the source of contamination and eliminate
the interference before processing samples. The LRB must contain ammonium acetate,
sodium omadine, and the internal standard at the same concentrations used to fortify all
field samples and CAL standards. Background from target analytes or contaminants that
interfere with the measurement of target analytes must be < 1/3 the MRL. If the target
analytes are detected in the LRB at concentrations equal to or greater than this level, then
536 -13
-------
all data for the problem analyte(s) must be considered invalid for all samples which
yielded a positive result.
NOTE: Although quantitative data below the MRL may not be accurate enough for data
reporting, such data are useful in determining the magnitude of background interference.
Therefore, blank contamination levels may be estimated by extrapolation when the
concentration is below the MRL.
9.3.2 CONTINUING CALIBRATION CHECK (CCC) - CCC standards are analyzed at the
beginning of each Analysis Batch, after every ten field samples and at the end of the
Analysis Batch. See Section 10.4 for concentration requirements and acceptance criteria.
9.3.3 LABORATORY FORTIFIED BLANK (LFB) - Since this method utilizes procedural
calibration standards, which are fortified reagent waters, there is no difference between
the LFB and the continuing calibration check standard. Consequently, the analysis of a
separate LFB is not required as part of the ongoing QC; however, the acronym LFB is
used for clarity in the IDC.
9.3.4 INTERNAL STANDARDS (IS) - The analyst must monitor the peak areas of the
internal standards in all injections of the Analysis Batch. The IS responses (as indicated
by peak areas) for any chromatographic run must not deviate by more than + 50% from
the average areas measured during the initial calibration for the internal standards. If the
IS area counts for a sample do not meet this criterion, inject a second aliquot of the
suspect sample as part of the same or a new Analysis Batch.
9.3.4.1 If the re-injected aliquot produces an acceptable internal standard response, report
results for that aliquot.
9.3.4.2 If the IS area counts of the re-injected aliquot still do not meet the IS criterion,
check the IS area of the most recent CCC. If the IS criterion is met in the CCC but
not the sample, report the sample results as suspect/matrix.
9.3.4.3 If the IS area criterion is not met in both the sample and the CCC, instrument
maintenance such as sample cone cleaning may be necessary. Perform the
appropriate instrument maintenance and then re-inject the sample in a subsequent
analytical batch.
9.3.5 LABORATORY FORTIFIED SAMPLE MATRIX (LFSM) - Analysis of a LFSM (Sect.
3.9) is required in each Analysis Batch. The LFSM is processed and analyzed exactly like
a sample, and its purpose is to determine whether the sample matrix contributes bias to
the analytical results. The background concentrations of the analytes in the sample
matrix must be determined in a separate aliquot and the measured values in the LFSM
corrected for background concentrations. If a variety of different sample matrices are
analyzed regularly, for example drinking water from groundwater and surface water
sources, performance data should be collected for each source.
9.3.5.1 Within each Analysis Batch, a minimum of one field sample is fortified as an LFSM
for every 20 samples analyzed. The LFSM is prepared by fortifying a sample with
an appropriate amount of the analyte PDS (Sect. 7.2.2.2). Select a spiking
536 -14
-------
concentration that is greater than or equal to the native background concentration, if
known. Selecting a duplicate aliquot of a sample that has already been analyzed
aids in the selection of an appropriate spiking level. If this is not possible, use
historical data and rotate through low, medium and high calibration concentrations
when selecting a fortifying concentration.
9.3.5.2 Calculate the percent recovery (%REC) using the equation:
C
where
A = measured concentration in the fortified sample,
B = measured concentration in the unfortified sample, and
C = fortification concentration.
9.3.5.3 Recoveries for samples fortified at concentrations near or at the MRL (within a
factor of two times the MRL concentration) should be within + 50% of the true
value. Recoveries for samples fortified at all other concentrations should be within +
30% of the true value. If the accuracy for any analyte falls outside the designated
range, and the laboratory performance for that analyte is shown to be in control in
the CCCs, the recovery is judged to be matrix biased. The result for that analyte in
the unfortified sample is labeled suspect/matrix to inform the data user that the
results are suspect due to matrix effects.
9.3.5.3.1 Field samples that have native analyte concentrations above the DL but below
the MRL and are fortified at concentrations at or near the MRL should be
corrected for the native levels in order to obtain meaningful %REC values.
This example and the LRB (Sect. 9.3.1) are the only permitted use of analyte
results below the MRL.
9.3.6 LABORATORY DUPLICATE OR LABORATORY FORTIFIED SAMPLE MATRIX
DUPLICATE (LD or LFSMD) - Within each Analysis Batch, a minimum of one
Laboratory Duplicate (LD) or Laboratory Fortified Sample Matrix Duplicate (LFSMD)
must be analyzed. Laboratory Duplicates check the precision associated with laboratory
procedures. If target analytes are not routinely observed in field samples, a LFSMD
should be analyzed rather than a LD.
9.3.6. 1 Calculate the relative percent difference (RPD) for duplicate measurements (LDi
and LD2) using the equation:
LD, -LD
(LDj+LDj/2
9.3.6.2 RPDs for Laboratory Duplicates should be < 30%. Greater variability may be
observed when Laboratory Duplicates have analyte concentrations that are within a
factor of 2 x MRL. At these concentrations, Laboratory Duplicates should have
536 -15
-------
RPDs that are < 50%. If the RPD of the analyte falls outside the designated range,
and the laboratory performance for the analyte is shown to be in control in the CCC,
the precision is judged to be matrix influenced. The result from the unfortified
sample is labeled suspect/matrix to inform the data user that the results are suspect
due to matrix effects.
9.3.6.3 If a LFSMD is analyzed instead of a Laboratory Duplicate, calculate the relative
percent difference (RPD) for duplicate LFSMs (LFSM and LFSMD) using the
equation:
LFSM-LFSMD
(LFSM + LFSMD)/2
9.3.6.4 RPDs for duplicate LFSMs should be < 30%. Greater variability may be observed
when fortified LFSMs have analyte concentrations that are within 2 x MRL.
LFSMs at these concentrations should have RPDs that are < 50%. If the RPD of the
analyte falls outside the designated range, and the laboratory performance for the
analyte is shown to be in control in the CCC, the precision is judged to be matrix
influenced. The result from the unfortified sample is labeled suspect/matrix to
inform the data user that the results are suspect due to matrix effects.
9.3.7 QUALITY CONTROL SAMPLE (QCS) - A QCS is required if an alternate commercial
source is available for the target analytes. A QCS should be evaluated as part of the IDC
(Sect. 9.2.5) and each time new PDS solutions are prepared. If standards are prepared
infrequently, analyze a QCS at least quarterly. The QCS should be fortified near the
midpoint of the calibration range and analyzed as a CCC. The acceptance criteria for the
QCS are the same as the mid- and high-level CCCs (Sect. 10.4). If measured analyte
concentrations are not of acceptable accuracy, check the entire analytical procedure to
locate and correct the problem.
9.4 METHOD MODIFICATION QC REQUIREMENTS - The analyst is permitted to modify
the separation technique, LC column, mobile phase composition, LC conditions and MS
conditions.
9.4.1 Each time method modifications are made, the analyst must repeat the procedures of the
IDC (Sect. 9.2) and verify that all QC criteria can be met in ongoing QC samples (Sect.
9.3).
9.4.2 The analyst is also required to evaluate and document method performance for the
proposed method modifications in real matrices that span the range of waters that the
laboratory analyzes. This additional step is required because modifications that perform
acceptably in the IDC, which is conducted in reagent water, can fail ongoing method QC
requirements in real matrices. This is particularly important for methods subject to
matrix effects, such as LC/MS-based methods. If, for example, the laboratory analyzes
finished waters from both surface and groundwater municipalities, this requirement can
be accomplished by assessing precision and accuracy (Sects. 9.2.2 and 9.2.3) in a surface
water with moderate to high total organic carbon ( e.g., 2 mg/L or greater) and a hard
groundwater (e.g., 250 mg/L as CaCOs equivalent, or greater).
536 -16
-------
9.4.3 The results of Sections 9.4.1 and 9.4.2 must be appropriately documented by the analyst
and should be independently assessed by the laboratory's QA officer prior to analyzing
field samples.
9.4.3.1 When implementing method modifications, it is the responsibility of the laboratory
to closely review the results of ongoing QC, and in particular, the results associated
with the LFSM (Sect. 9.3.5), LFSMD (Sect. 9.3.6), CCCs (Sect. 9.3.2), and the IS
area counts (Sect. 9.3.4). If repeated failures are noted, the modification must be
abandoned.
10 CALIBRATION AND STANDARDIZATION
10.1 Demonstration and documentation of acceptable MS calibration and initial analyte
calibration are required before performing the IDC (Sect. 9.2) and prior to analyzing field
samples. After initial calibration is successful, Continuing Calibration Check (CCC)
standards are required at the beginning and the end of each analysis batch, and after every
tenth sample (Sect. 10.4). Verification of the MS calibration and the initial calibration
should be repeated each time a major instrument modification or maintenance is performed.
10.2 LC/MS/MS CALIBRATION AND OPTIMIZATION
10.2.1 MASS CALIBRATION - Calibrate the mass spectrometer with the calibration
compounds and procedures specified by the manufacturer.
10.2.2 OPTIMIZING MS PARAMETERS - This method requires the utilization of ESI and
Multiple Reaction Monitoring as defined above (Sect. 3.15). LC/MS instruments have
several instrumental parameters that must be optimized. Each LC/MS system will have
different optimal conditions which are influenced by the source geometry and system
design. Due to the differences in design, the recommendations of the instrument
manufacturer should be followed when tuning the instrument. In the development of this
method, instrumental parameters were optimized for the precursor and product ions listed
in Section 17, Table 3. Literature resources to date describe similar precursor and
product ions.6'7 Product ions other than those listed may be selected; however, the
analyst is cautioned to avoid using ions with lower mass and/or common ions that may
not provide sufficient discrimination between the analytes of interest and co-eluting
interferences.
10.2.2.1 Prepare 0.05 - 1.0 |ig/mL solutions of the target analytes in the approximate mobile
phase compositions that correspond to the elution mobile phase compositions. This
step will need to be repeated and the parameters re-optimized if an analyte is
prepared in a mobile phase composition that is significantly different from the
mobile phase composition at the time the analyte elutes from the column.
10.2.2.2 Optimize the response of the precursor ion (Sect. 3.16) for each target analyte by
infusing the prepared solutions (Sect. 10.2.2.1) directly into the mass spectrometer
as recommended by the instrument manufacturer. The LC should be operating with
a mobile phase composition and flow rate comparable to the elution conditions.
Vary the MS parameters (source voltages, source and desolvation temperatures, gas
536 -17
-------
flows, etc.) until optimal analyte responses are determined. The target analytes may
have different optima, thus requiring some compromise. LC/MS systems are
subject to drift, which can affect run-to-run precision. Accordingly, the optimal
conditions may be those that yield good response while minimizing the potential for
drift. See Section 17, Table 2 for the ESI+ MS conditions used during method
development.
10.2.2.3 In all cases, a precursor ion must be isolated (e.g., in a quadrupole or an ion trap)
prior to fragmentation. Optimize the response of the product ion (Section 3.18) for
each target by infusing the prepared solutions (Sect. 10.2.2.1) directly into the mass
spectrometer as recommended by the instrument manufacturer. The LC should be
operating with a mobile phase composition and flow rate comparable to the elution
conditions. The MS/MS parameters (collision gas pressure, collision energy, etc.)
are varied until optimal analyte responses are determined. The triazine compounds
listed in this method exhibited very similar optimal MS/MS conditions. See Section
17, Table 3 for the MS/MS conditions used in method development.
10.2.3 LC INSTRUMENT CONDITIONS - LC operating conditions are described in Section
17, Table 1. Conditions different from those described (e.g., LC columns and mobile
phases) may be used if the QC criteria in Sections 9.2 and 9.3 are met and
chromatographic separation of the target analytes is achieved (resolution > 50%).
NOTE: Chromatographic separation as defined does not include the isotopically
enriched internal standards, which are mass separated. Co-elution of the internal
standards with their analogous target analytes helps mitigate matrix suppression and/or
enhancement effects.8
10.2.4 ESTABLISH LC/MS RETENTION TIMES AND MRM SEGMENTS - Inject a mid- to
high-level CAL standard under optimized LC/MS conditions to obtain the retention times
of each target analyte. Divide the chromatogram into segments that contain one or more
chromatographic peaks. For maximum sensitivity in subsequent MS/MS analyses,
minimize the number of MRM (Sect. 3.15) transitions that are simultaneously monitored
within each segment.
10.3 INITIAL CALIBRATION - During method development, daily calibrations were performed;
however, it is permissible to verify the calibration with daily CCCs. Calibration must be
performed using peak areas and the internal standard technique. Calibration using peak
heights or external calibration is not permitted.
10.3.1 CALIBRATION STANDARDS - Prepare a set of at least five CAL standards as
described in Section 7.2.3. The lowest concentration of the calibration standards must be
at or below the MRL. The MRL must be confirmed using the procedure outlined in
Section 9.2.4 after establishing the initial calibration. Additionally, field samples must be
quantified using a calibration curve that spans the same concentration range used to
collect the IDC data (Sect. 9.2).
10.3.2 CALIBRATION - The LC/MS system is calibrated using peak areas and the internal
standard technique. Concentrations may be calculated using the data system software to
generate a linear regression or quadratic calibration curve. Forcing the calibration curve
536 -18
-------
through the origin is not recommended. The MS/MS instrument used during method
development was calibrated using weighted (1/X) quadratic curves. Data may be fit with
either a linear regression (response vs. concentration) or quadratic fit (response vs.
concentration) and weighting may be used, if necessary.
10.3.3 CALIBRATION ACCEPTANCE CRITERIA - The validation of the calibration is
determined by calculating the concentration of the analyte from each of the analyses used
to generate the calibration curve. Calibration points that are < MRL should calculate to
be within + 50% of their true value. All other calibration points should calculate to be
within + 30% of their true value. If these criteria cannot be met, the analyst will have
difficulty meeting ongoing QC criteria. In this case, corrective action should be taken to
reanalyze the calibration standards, restrict the range of calibration, or select an alternate
method of calibration.
10.4 CONTINUING CALIBRATION CHECKS (CCCs) - The CCC verifies the calibration at the
beginning, after every tenth field sample and at the end of each Analysis Batch. CCCs are
not counted as samples. The beginning CCC for each Analysis Batch must be at or
below the MRL. This CCC verifies instrument sensitivity prior to the analysis of samples.
Subsequent CCCs should alternate between a medium and high concentration CAL standard.
10.4.1 Inject an aliquot of the appropriate concentration CAL standard and analyze with the
same conditions used during the initial calibration.
10.4.2 Determine that the absolute areas of the quantitation ions of each of the internal standards
have not changed by more than + 50% from the average areas measured during the initial
calibration. If IS areas have changed by more than + 50%, remedial action is necessary
(Sect. 10.4.4). Recalibrate according to Section 10.3 and verify sensitivity by analyzing a
CCC at or below the MRL.
10.4.3 Calculate the concentration of each analyte in the CCC. The CCC fortified at < MRL
must calculate to be within + 50% of its true value. CCCs fortified at all other levels
must calculate to be within + 30%. If these conditions do not exist, then all data for the
analyte must be considered invalid, and remedial action (Sect. 10.4.4) must be taken.
The remedial action may require recalibration. Any field samples that have been
analyzed since the last acceptable CCC that are still within holding time should be
reanalyzed after adequate calibration has been restored.
10.4.4 REMEDIAL ACTION - Failure to meet CCC QC performance criteria requires remedial
action. Maintenance such as cleaning an ion source probe, sample cone, ion lenses
and/or replacing LC columns may be required. Following major maintenance, the analyst
must return to the initial calibration step (Sect. 10.3).
11. PROCEDURE
11.1 Important aspects of this analytical procedure include proper sample collection and storage
(Section 8), ensuring that the instrument is properly calibrated (Section 10) and that all
required QC are met (Section 9). This section describes the procedures for sample
preparation and analysis.
536 -19
-------
11.2 SAMPLE PREPARATION
11.2.1 Samples are dechlorinated, preserved, collected and stored as described in Section 8.1.
All field and QC samples must contain the dechlorinating and preservation agents listed
in Section 8.1.1, including the LRB. Allow field samples to reach room temperature
prior to analysis.
11.2.2 Transfer a 1.0 mL aliquot of each field or QC sample to an autosampler vial; as
appropriate, samples that will be chromatographed using small particle diameter LC
columns (< 2 jim) may be filtered prior to analysis (see note below). Add the IS PDS
(Sect. 7.2.1.2), cap and mix well. Larger amounts of sample may be used, but the amount
of internal standard must be increased to give a final concentration that is equal to the
CAL IS concentration.
NOTE: If filtering is incorporated as part of the sample preparation, the first lot of filters
must be subjected to the procedures outlined in the IDC (Sect. 9.2) and meet the
acceptance criteria defined in Table 9 to ensure that they do not introduce interferences or
retain any of the target analytes. Verification of subsequent lots of filters can be
accomplished by examining a LRB and duplicate samples of LFBs fortified at the MRL.
The LFBs should calculate to be within + 50% of the true value. If the LRB or the LFBs
fail this evaluation, the full IDC will need to be repeated with the new lot of filters. CAL
standards and CCCs should not be filtered in order to identify potential losses associated
with the sample filtration devices.
11.3 SAMPLE ANALYSIS
11.3.1 Establish MS operating conditions equivalent to those summarized in Tables 1-3 of
Section 17 as per the guidance in Section 10.2. Column choice and instrument
parameters should be optimized prior to initiation of the IDC.
11.3.2 Establish an appropriate retention time window (Sect. 10.2.4) for each analyte to identify
them in QC and field sample chromatograms. This should be based on measurements of
actual retention time variation for each compound in standard solutions analyzed on the
LC over the course of time. The suggested variation is plus or minus three times the
standard deviation of the retention time for each compound for a series of injections. The
injections from the initial calibration and from the IDC (Sect. 9.2) may be used to
calculate a suggested window size. However, the experience of the analyst should weigh
heavily on the determination of an appropriate retention window size.
11.3.3 Establish a valid initial calibration following the procedures outlined in Section 10.3 or
confirm that the calibration is still valid by running a CCC as described in Section 10.4.
If establishing an initial calibration for the first time, complete the IDC as described in
Section 9.2 prior to analyzing field samples.
11.3.4 Analyze aliquots of field and QC samples at appropriate frequencies (Section 9) with the
LC/MS/MS conditions used to acquire the initial calibration.
536 -20
-------
11.3.5 At the conclusion of data acquisition, use the same software settings that were used in the
calibration procedure to identify peaks of interest in the predetermined retention time
windows. Identify an analyte by comparison of its retention time with that of the
corresponding analyte peak in a CAL standard or CCC.
11.3.6 EXCEEDING CALIBRATION RANGE - The analyst must not extrapolate beyond the
established calibration range. If an analyte result exceeds the range of the initial
calibration curve, the sample may be diluted using reagent water containing 20 mM
ammonium acetate and 64 |ig/L sodium omadine and with the appropriate amount of
internal standard added to match the original level. Re-inject the diluted sample.
Incorporate the dilution factor into final concentration calculations. The resulting data
should be documented as a dilution and the dilution will affect analyte MRLs.
12. DATA ANALYSIS AND CALCULATIONS
12.1 While complete chromatographic resolution is not necessary for accurate and precise
measurements of analyte concentrations using MS/MS in reagent water, it is important in real
matrices. The analyst must avoid chromatographic modifications that result in suppression
or enhancement of analyte responses due to co-eluting matrix components. This is especially
a problem with samples that contain higher TOC levels (Sect. 13.4.2). In validating this
method, concentrations were calculated using the product ions listed in Table 3 of Section 17.
Other ions may be selected at the discretion of the analyst as discussed in Section 10.2.2.
12.2 Calculate analyte concentrations using the multipoint calibration established in Section 10.3.
Report only those values that fall between the MRL and the highest CAL standard. Samples
with target analyte responses that exceed the highest CAL standard require dilution and
reanalysis (Sect. 11.3.6).
12.3 Calculations must use all available digits of precision, but final reported concentrations
should be rounded to an appropriate number of significant figures (one digit of uncertainty),
typically two, and not more than three significant figures.
12.4 Prior to reporting the data, the chromatograms should be reviewed for any incorrect peak
identifications or poor integrations.
12.5 Prior to reporting data, the laboratory is responsible for assuring that QC requirements have
been met or that any appropriate qualifier is documented.
13. METHOD PERFORMANCE
13.1 PRECISION, ACCURACY AND DETECTION LIMITS - Tables for these data are
presented in Section 17. Instrumental conditions are presented in Tables 1 through 3. DLs
and LCMRLs are presented in Table 4. LCMRLs were determined and calculated using a
procedure described elsewhere.1 Single laboratory precision and accuracy data are presented
for three water matrices: reagent water (Table 5); chlorinated (finished) surface water (Table
6); and chlorinated (finished) groundwater (Table 7). Figure 1 shows an overlaid
chromatogram of the normalized MS/MS responses for the analytes obtained under the
conditions employed during method development.
536 -21
-------
13.2 SECOND LABORATORY EVALUATION - The performance of this method was
demonstrated by a second laboratory, with results similar to those reported in Section 17.
The authors wish to acknowledge the work of Dr. Jody Shoemaker of the U.S. EPA National
Exposure Research Laboratory, Cincinnati, OH for conducting the second laboratory
evaluation.
13.3 ANALYTE STABILITY STUDIES - Chlorinated surface water samples, fortified with
target analytes at 5 |ig/L, were preserved as required in Section 8 and stored over a 28-day
period. The accuracy and precision of three replicate analyses, conducted on day 0, 7, 14, 21
and 28, are presented in Section 17, Table 8. These data were used to determine the 28-day
holding time.
13.4 PROBLEM COMPOUNDS - During method development and the second laboratory
evaluation, the following problems were observed for Atrazine-desethyl-desisopropyl
(DACT):
13.4.1 DACT is subject to hydrolysis at both high (> 9) and low (< 5) pH but has acceptable
stability in neutral aqueous solutions. DACT is also degraded in the presence of free
available chlorine, but stable in the presence of chloramines. Ammonium acetate
mitigates both of these modes of loss and has sufficient buffer capacity at 20 mM to
buffer finished groundwaters with relatively high pH that also have buffer capacity.
13.4.2 DACT and its internal standard (DACT-13C) exhibited ESI matrix suppression in the
second laboratory evaluation after being fortified into a chlorinated surface water
containing high total organic carbon (TOC) levels (9.5 mg/L). The DACT-13C response
was suppressed 26% relative to the calibration; however, since DACT was affected by
the suppression in an equivalent manner, the DACT recovery was 99%. The matrix
suppression occurred due to co-eluting TOC in the early part of the chromatographic run
and DACT is the analyte most susceptible to matrix effects. As discussed in Section
4.4.1, good performance is achieved by optimizing chromatographic separation, and in
particular, the separation of DACT from early eluting contaminants. Modifications that
seek to shorten the chromatographic separation without accounting for the need to
separate the analytes from early eluting interferences may result in failure to meet QC
criteria.
14. POLLUTION PREVENTION
14.1 For information about pollution prevention that may be applicable to laboratory operations,
consult "Less is Better: Laboratory Chemical Management for Waste Reduction" available
from the American Chemical Society's Department of Government Relations and Science
Policy, 1155 16th Street N.W., Washington, D.C., 20036, or on-line at
http://www.ups.edu/community/storeroom/Chemical_Wastes/wastearticles.htm.
15. WASTE MANAGEMENT
15.1 The analytical procedures described in this method generate relatively small amounts of
waste since only small amounts of reagents and solvents are used. The matrices of concern
are finished drinking water or source water. However, the Agency requires that laboratory
waste management practices be conducted consistent with all applicable rules and
536 -22
-------
regulations, and that laboratories protect the air, water, and land by minimizing and
controlling all releases from fume hoods and bench operations. Also, compliance is required
with any sewage discharge permits and regulations, particularly the hazardous waste
identification rules and land disposal restrictions. For further information on waste
management, see the publications of the American Chemical Society's Laboratory
Environment, Health & Safety Task Force on the Internet at
http://membership.acs.0rg/c/ccs/publications.htm. Additional waste management
information can be found in "Laboratory Waste Minimization and Pollution Prevention,"
Copyright © 1996 Battelle Seattle Research Center, which can be located at
http://www.p2pays.org/ref/01/text/00779/index2.htm.
16. REFERENCES
1. Winslow, S.D.; Pepich, B.V.; Martin, J.J.; Hallberg, G.R.; Munch D.J.; Frebis, C.P.; Hedrick, E.J.;
Krop, R.A. Statistical Procedures for Determination and Verification of Minimum Reporting Levels
for Drinking Water Methods. Environ. Sci. Technol. 2006, 40, 281-288.
2. Glaser, J.A.; Foerst, D.L.; McKee, G.D.; Quave, S.A.; Budde, W.L. Trace Analyses for
Wastewaters. Environ. Sci. Technol. 1981, 75, 1426-1435.
3. "Guidelines, Recommendations, and Regulations for Handling Antineoplastic Agents", Center for
Disease Control, National Institute for Occupational Safety and Health,
http://www.cdc.gov/niosh/topics/antineoplastic/pubs.html#b.
4. "OSHA Safety and Health Standards, General Industry," (29CFR1910), Occupational Safety and
Health Administration, OSHA 2206 (Revised, January 1976).
5. "Safety in Academic Chemistry Laboratories," American Chemical Society Publication, Committee
on Chemical Safety, 3rd Edition, 1979.
6. Steen, Ruud J.C.A.; Hogenboom, Ariadne C.; Leonards, Pirn E.G.; Peerboom, Renee A.L.; Cofmo,
Wim P.; Brinkman, Udo A.Th. Ultra-trace Level Determination of Polar Pesticides and Their
Transformation Products in Surface and Estuarine Water Samples Using Column Liquid
Chromatography-Electrospray Tandem Mass Spectrometry. J. Chromatogr. A 1999, 557, 157-166.
7. Huang, S.; Mayer, T.J.; Yokley, R.A.; Perez, R. Direct Aqueous Injection Liquid
Chromatography/Electrospray lonizati on-Mass Spectrometry/Mass Spectrometry Analysis of Water
for Atrazine, Simazine, and their Chlorotriazine Metabolites. J. Agric. Food Chem. 2006, 54, 713-
719.
8. Kloepfer, A.; Quintana, J.B.; Reemtsma, T. Operational Options to Reduce Matrix Effects in Liquid
Chromatrography-Electrospray lonizati on-Mass Spectrometry Analysis of Aqueous Environmental
Samples. J. Chromatogr. A 2005, 1067, 153-160.
536 -23
-------
17. TABLES, DIAGRAMS, FLOWCHARTS, AND VALIDATION DATA
TABLE 1. LC CONDITIONS a
Time (min.)
Initial
5.0
25.0
25.1
35.0
35.1
45.0
45.1
65.0
%5 mM NH4OAc
90
90
70
60
60
50
50
90
90
% Methanol
10
10
30
40
40
50
50
10
10
Chromatographic conditions: Column: Waters Xterra® RP18, 2.1 x 150 mm, 3.5 |j,m particle
diameter; Injection Volume: 100 |j,L; Flow Rate: 0.250 mL/min.
TABLE 2. ESI-MS/MS METHOD CONDITIONS
MS Parameter
Polarity
Capillary Voltage
Source Temperature
Desolvation Temperature
Desolvation Gas Flow
Cone Gas Flow
Collision Cell Pressure
Collision Gas
Setting
Positive Ion Mode
0.5 kV
100°C
260 °C
616L/hr
26 L/hr
2.1 e-3 mbar
Argon
536 -24
-------
TABLE 3. LC/ESI-MS/MS RETENTION TIMES, PRECURSOR AND PRODUCT IONS,
AND COLLISION ENERGIES
MRM
Segment
#
1
1
2
2
3
3
4
4
4
4
5
5
6
6
Analyte
Atrazine-desethyl-desisopropyl-13C (IS)
Atrazine-desethyl-desisopropyl
Atrazine-desisopropyl-d5 (IS)
Atrazine-desisopropyl
Atrazine-desethyl-d7 (IS)
Atrazine-desethyl
Simazine-dlO (IS)
Simazine
Cyanazine-d5 (IS)
Cyanazine
Atrazine-d5 (IS)
Atrazine
Propazine-dl4 (IS)
Propazine
Ret.
Time
(min.)
3.78
3.78
11.26
11.57
18.66
19.09
30.48
30.96
32.03
32.20
37.03
37.37
42.87
43.51
Precursor
Ion
(m/z)
149
146
179
174
195
188
212
202
246
241
221
216
244
230
Product
Ion
(m/z)
105.5
103.5
137
132
147
146
137
132
219
214
179
174
196
188
Collision
Energy
(eV)
22
22
22
22
22
22
20
20
20
20
23
23
20
20
TABLE 4. LC/ESI-MS/MS LOWEST CONCENTRATION MRLs AND DLs a
Analyte
Atrazine-desethyl-desisopropyl
Atrazine-desisopropyl
Atrazine-desethyl
Simazine
Cyanazine
Atrazine
Propazine
LCMRL Concentration
Levels (ug/L)
0.104,0.208,0.260,0.520
0.102,0.204,0.255,0.510
0.021,0.052,0.104,0.130
0.020,0.050,0.100,0.125
0.019,0.048,0.096,0.120
0.021,0.052,0.104,0.130
0.020,0.050,0.100,0.125
Calculated
LCMRL (ng/L)
0.14
0.11
0.026
0.020
0.027
0.037
0.028
Calculated
DL (ng/L)
0.028
0.043
0.017
0.010
0.010
0.010
0.010
DLs calculated using data acquired on a single day at the lowest LCMRL concentration level (n=7 sample
replicates).
536 -25
-------
TABLE 5. LC/ESI-MS/MS PRECISION AND ACCURACY OF METHOD ANALYTES
FORTIFIED AT 0.5 ug/L AND 5.0 ug/L IN REAGENT WATER (n=7 SAMPLES)
Analyte
Atrazine-desethyl-desisopropyl
Atrazine-desisopropyl
Atrazine-desethyl
Simazine
Cyanazine
Atrazine
Propazine
Fortified Cone. = 0.5 ug/L
(n=7)
Mean %
Recovery
98.9
100
99.4
101
97.7
99.4
103
Relative Standard
Deviation (%RSD)
4.9
3.4
2.7
3.0
1.3
2.5
2.9
Fortified Cone. = 5.0 ug/L
(n=7)
Mean %
Recovery
89.4
96.1
97.6
101
99.3
99.7
99.0
Relative Standard
Deviation
(%RSD)
4.2
3.7
1.3
1.9
1.4
1.6
2.0
TABLE 6. LC/ESI-MS/MS PRECISION AND ACCURACY OF METHOD ANALYTES
FORTIFIED AT 0.5 ug/L AND 5.0 ug/L IN A CHLORINATED SURFACE
WATER a (n=7 SAMPLES)
Analyte
Atrazine-desethyl-desisopropyl
Atrazine-desisopropyl
Atrazine-desethyl
Simazine
Cyanazine
Atrazine
Propazine
Fortified Cone. = 0.5 ug/L
(n=7)
Mean %
Recovery b
102
100
95.5
94.5
95.3
97.4
94.9
Relative Standard
Deviation (%RSD)
8.4
6.6
1.5
3.7
4.2
6.1
6.0
Fortified Cone. = 5.0 ug/L
(n=7)
Mean %
Recovery11
103
101
97.9
99.4
99.4
101
105
Relative Standard
Deviation
(%RSD)
2.7
3.7
0.9
1.7
1.7
0.9
1.5
a Surface water physical parameters: pH=7.1; Hardness = 154 mg/L; Free chlorine =1.5 mg/L (Hach).
b Recoveries corrected for native levels in the unfortified matrix: Atrazine-desethyl-desisopropyl (0.043 (ig/L),
Atrazine-desisopropyl (0.001 (ig/L), Atrazine-desethyl (0.031 (ig/L), Simazine (0.054 (ig/L), Cyanazine (0.018
(ig/L), Atrazine (0.316 (ig/L) and Propazine (0.038 (ig/L).
536 -26
-------
TABLE 7. LC/ESI-MS/MS PRECISION AND ACCURACY OF METHOD ANALYTES
FORTIFIED AT 0.5 ug/L AND 5.0 ug/L IN A CHLORINATED
GROUNDWATER a (n=7 SAMPLES)
Analyte
Atrazine-desethyl-desisopropyl
Atrazine-desisopropyl
Atrazine-desethyl
Simazine
Cyanazine
Atrazine
Propazine
Cone. = 0.5 jig/L
(n=7)
Mean %
Recovery b
106
99.6
95.7
94.6
93.5
96.7
91.7
Relative Standard
Deviation
(%RSD)
3.5
5.2
1.7
3.1
1.8
0.9
3.7
Cone. = 5.0 ug/L
(n=7)
Mean %
Recovery11
95.7
103
98.6
98.0
101
100
99.7
Relative Standard
Deviation (%RSD)
2.5
3.0
1.5
1.5
1.3
1.5
1.4
a Groundwater physical parameters: pH = 7.5; Hardness = 360 mg/L; Free chlorine = 0.49 mg/L.
b Recoveries corrected for native levels in the unfortified matrix: Atrazine-desethyl-desisopropyl (0.032(ig/L),
Atrazine-desisopropyl (not detected), Atrazine-desethyl (0.036 (ig/L), Simazine (0.030 (ig/L), Cyanazine (0.018
(ig/L), Atrazine (0.057 (ig/L), and Propazine (0.036 (ig/L).
536 -27
-------
TABLE 8. LC/ESI-MS/MS SAMPLE HOLDING TIME DATA FOR METHOD ANALYTES FORTIFIED AT 5.0 jig/L IN A
CHLORINATED SURFACE WATER a (n=3 SAMPLES)
Analyte
Atrazine
Atrazine-desethyl
Atrazine-desethyl-desisopropyl
Atrazine-desisopropyl
Cyanazine
Propazine
Simazine
DayO
%Rec
98.0
107
95.9
105
99.3
112
94.2
%RSD
4.1
2.5
7.7
3.5
4.7
2.9
6.0
Day 7
%Rec
97.4
104
93.9
89.8
94.0
102
99.6
%RSD
2.9
4.1
6.2
7.6
1.2
4.8
4.8
Day 14
%Rec
99.4
111
102
95.4
94.8
109
93.6
%RSD
1.1
1.7
1.2
1.1
1.6
2.5
6.8
Day 21
%Rec
94.3
103
90.8
93.5
87.8
103
89.1
%RSD
7.0
4.7
6.6
7.9
4.7
6.1
11.8
Day 28
%Rec
99.3
108
103
102
89.5
109
100
%RSD
0.8
1.5
3.6
1.8
0.5
1.6
2.5
Surface water physical parameters: pH = 7.0; Hardness=154 mg/L; Free chlorine = 1.2 mg/L (Hach).
536 -28
-------
TABLE 9. INITIAL DEMONSTRATION OF CAPABILITY (IDC) QUALITY CONTROL
REQUIREMENTS
Method
Reference
Section
9.2.1
Section
9.2.1
Section
9.2.2
Section
9.2.3
Section
9.2.4
Section
9.2.5
Requirement
Demonstration of
Low System
Background
Test For System
Carryover
Demonstration of
Precision
Demonstration of
Accuracy
Minimum
Reporting Limit
(MRL)
Confirmation
Quality Control
Sample (QCS)
Specification and
Frequency
Analyze a LRB prior to
any other IDC steps.
Analyze a reagent water
blank after the high CAL
standard during the IDC
calibration.
Analyze 7 replicate CCCs
fortified near the midrange
concentration.
Calculate average recovery
for replicates used in
Section 9.2.2.
Fortify and analyze 7
replicate CCCs at the
proposed MRL
concentration. Calculate
the mean and the Half
Range (HR). Confirm that
the Upper PIR and Lower
PIR (Sect. 9.2.4.2) meet
the recovery criteria.
Analyze mid-level QCS
sample.
Acceptance Criteria
Demonstrate that all target
analytes are < 1/3 of the MRL
and that possible interferences
from reagents and glassware
do not prevent the
identification and quantitation
of target analytes.
Demonstrate that the target
analytes are < 1/3 of the MRL
and that carry-over from
previous samples does not
prevent the identification and
quantitation of target analytes.
%RSD must be < 20%
Mean recovery within + 30%
of the true value.
Upper PIR < 150%
Lower PIR > 50%
Results must be within + 30%
of the true value.
536 -29
-------
TABLE 10. ONGOING QUALITY CONTROL REQUIREMENTS
Method
Reference
Requirement
Specification and
Frequency
Acceptance Criteria
Section
8.3
Sample Holding
Time
28 days when processed and
stored according to Sections
8.1 and 8.2.
Sample results are valid only if
samples are analyzed within the
sample holding time.
Section
10.3
Initial Calibration
Use the internal standard
calibration technique to
generate a linear or quadratic
calibration curve. Use at
least 5 standard
concentrations. Validate the
calibration curve as
described in Section 10.3.3.
When each calibration standard
is calculated as an unknown
using the calibration curve, the
lowest level standard should be
within + 50% of the true value.
All other points should be
within + 30% of the true value.
Section
9.3.1
Laboratory Reagent
Blank (LRB)
Daily, or with each Analysis
Batch of up to 20 field
samples, whichever is more
frequent.
Demonstrate that all target
analytes are below 1/3 the
MRL, and that possible
interference from reagents and
glassware do not prevent
identification and quantitation
of target analytes.
Section
10.4
Continuing
Calibration Check
(CCC)
Verify initial calibration by
analyzing a low-level CCC at
the beginning of each
Analysis Batch. Subsequent
CCCs are required after
every 10 field samples, and
after the last field sample in a
batch.
The lowest level CCC must be
within + 50% of the true value.
All other points must be within
+ 30% of the true value.
Results for field samples that
are not bracketed by acceptable
CCCs are invalid.
536 -30
-------
TABLE 10. ONGOING QUALITY CONTROL REQUIREMENTS (CONTINUED)
Method
Reference
Requirement
Specification and
Frequency
Acceptance Criteria
Section
9.3.4
Internal Standard
(IS)
Isotopically labeled internal
standards are added to all
standards and samples.
Peak area counts for each IS
must be within + 50% of the
average peak areas in the initial
calibration.
Section
9.3.5
Laboratory
Fortified Sample
Matrix (LFSM)
Analyze one LFSM per
Analysis Batch. Fortify the
LFSM with target analytes at
a concentration close to but
greater than the native
concentrations (if known).
Calculate LFSM recoveries.
For LFSMs fortified at
concentrations < MRL, the
result should be within + 50%
of the true value. All other
LFSMs should be within + 30%
of the true value.
Section
9.3.6
Laboratory
Fortified Sample
Matrix Duplicate
(LFSMD) or
Laboratory
Duplicate(LD)
Analyze at least one LFSMD
or LD with each Analysis
Batch of up to 20 field
samples.
For LFSMDs or LDs, RPDs
should be < 30%.
Sections
9.3.7
Quality Control
Sample (QCS)
Analyze mid-level QCS
sample when new PDS
solutions are prepared, or at
least quarterly if PDS
solutions are prepared less
frequently.
Results must be + 30% of the
true value.
536 -31
-------
FIGURE 1. NORMALIZED, OVERLAID CHROMATOGRAMS OF LC/ESI-MS/MS TRANSITIONS FOR METHOD 536
ANALYTES FORTIFIED AT 0.5 jig/L IN REAGENT WATER a
100-1
H
O
o-
0.00
w
Q
c<3
yanaz
Atrazme
C
------- |