EPA815-R-05-009
METHOD 314.1 DETERMINATION OF PERCHLORATE IN DRINKING WATER USING
INLINE COLUMN CONCENTRATION/MATRIX ELIMINATION ION
CHROMATOGRAPHY WITH SUPPRESSED CONDUCTIVITY
DETECTION
Revision 1.0
May 2005
Herbert P. Wagner (Lakeshore Engineering Services, Inc.)
Barry V. Pepich (Shaw Environmental, Inc.)
Chris Pohl, Douglas Later, Robert Joyce, Kannan Srinivasan, Brian DeBorba, Dave Thomas, and
Andy Woodruff (Dionex, Inc., Sunnyvale, CA)
David J. Munch (U.S. EPA, Office of Ground Water and Drinking Water)
TECHNICAL SUPPORT CENTER
OFFICE OF GROUND WATER AND DRINKING WATER
U. S. ENVIRONMENTAL PROTECTION AGENCY
CINCINNATI, OHIO 45268
314.1-1
-------
EPA815-R-05-009
METHOD 314.1
DETERMINATION OF PERCHLORATE IN DRINKING WATER BY INLINE COLUMN
CONCENTRATION/MATRIX ELIMINATION ION CHROMATOGRAPHY WITH
SUPPRESSED CONDUCTIVITY DETECTION
1. SCOPE AND APPLICATION
1.1 This is a sample pre-concentration, matrix elimination ion chromatographic (1C) method using
suppressed conductivity detection for the determination of perchlorate in raw and finished
drinking waters. This method requires the use of a confirmation column to validate all
perchlorate concentrations reported at or above the MRL on the primary column. Precision
and accuracy data have been generated for perchlorate, with both the primary and
confirmation columns, in reagent water, finished groundwater, surface water and a Laboratory
Fortified Synthetic Sample Matrix (LFSSM). The single laboratory Lowest Concentration
Minimum Reporting Level (LCMRL) has also been determined in reagent water.l
Chemical Abstract Services
Analvte Registry Number (CASRN)
Perchlorate 14797-73-0
1.2 The Minimum Reporting Level (MRL) is the lowest analyte concentration that meets Data
Quality Objectives (DQOs) that are developed based on the intended use of this method. The
single laboratory LCMRL is the lowest true concentration for which the future recovery is
predicted to fall between 50 and 150 percent recovery with 99% confidence. The single
laboratory LCRML for perchlorate was 0.140 and 0.130 ug/L for the AS16 and AS20
columns, respectively. The procedure used to determine the LCMRL is described elsewhere.l
1.3 Laboratories using this method will not be required to determine the LCMRL, but will need to
demonstrate that their laboratory MRL for this method meets the requirements described in
Section 9.2.4.
1.4 Detection limit (DL) is defined as the statistically calculated minimum concentration that can
be measured with 99% confidence that the reported value is greater than zero.2 The DL for
perchlorate is dependent on sample matrix, fortification concentration, and instrument
performance. Determining the DL for perchlorate in this method is optional (Sect. 9.2.7).
The reagent water DL for the perchlorate was calculated to be 0.03 ug/L using 7 replicates of
a 0.10 |ig/L fortification level with the AS 16 columns and 0.03 ug/L for the AS20 columns.
These values are also provided in.-
1.5 This method is intended for use by analysts skilled in the operation of 1C instrumentation, and
the interpretation of the associated data.
314.1-2
-------
EPA815-R-05-009
2. SUMMARY OF METHOD
2.1 Water samples are collected in the field using a sterile filtration technique. The sample,
without cleanup, is concentrated onto the concentrator/trap column, which is placed in the
sample loop position and binds perchlorate more strongly than other matrix anions. The
sample matrix anions are rinsed from the concentrator column with 1 mL of 10 mM NaOH.
This weak rinse solution allows the concentrator to retain the perchlorate while eluting the
majority of the matrix anions, which are directed to waste. The concentrator column is
switched in-line and the perchlorate is eluted from the concentrator column with a 0.50 mM
NaOH solution. Following elution from the concentrator, the perchlorate is refocused onto
the front of the guard column. The eluent strength is then increased to 65 mM NaOH which
elutes the perchlorate from the guard column and onto the analytical column where
perchlorate is separated from other anions and remaining background interferences. The
sample loading and matrix elimination steps must use the same eluent flow direction as the
elution and analytical separation steps. Perchlorate is subsequently detected using suppressed
conductivity and is quantified using an external standard technique. Confirmation of any
perchlorate concentration reported at or above the MRL on the primary column is
accomplished with a second analytical column that has a dissimilar separation mechanism.
3. DEFINITIONS
3.1 ANALYSIS BATCH - A sequence of field samples, which are analyzed within a 30-hour
period and include no more than 20 field samples. An Analysis Batch must also include all
required QC samples, which do not contribute to the maximum field sample total of 20. For
this method, the required QC samples include:
Laboratory Synthetic Sample Matrix Blank (LSSMB)
Continuing Calibration Check (CCC)
Laboratory Fortified Synthetic Sample Matrix (LFSSM) CCC Standards
Laboratory Fortified Sample Matrix (LFSM)
Laboratory Duplicate (LD) or a Laboratory Fortified Sample Matrix Duplicate
(LFSMD).
3.2 ANALYTE FORTIFICATION SOLUTIONS (AFS) - The Analyte Fortification Solutions
are prepared by dilution of the Analyte Secondary Dilution Solutions (SDS) and are used to
fortify the LFSMs and the LFSMDs with perchlorate. It is recommended that multiple
concentrations be prepared so that the fortification levels can be rotated or adjusted to the
concentration of target analyte in the native samples.
3.3 CALIBRATION BLANK (CB) - An aliquot of reagent water or other blank matrix that is
treated exactly as a CCC. The CB is not sterile filtered and is used to determine if the
method analyte or other interferences are present in the laboratory environment, the reagents,
or the apparatus during the IDC calibration.
3.4 CALIBRATION STANDARD (CAL) - A solution of the target analyte prepared from the
Perchlorate Primary Dilution Solution or Perchlorate Stock Standard Solution. The CAL
solutions are used to calibrate the instrument response with respect to analyte concentration.
314.1-3
-------
EPA815-R-05-009
3.5 CONTINUING CALIBRATION CHECK STANDARD (CCC) - A calibration check
standard containing the method analyte which is analyzed periodically throughout an
Analysis Batch, to verify the accuracy of the existing calibration for that analyte.
3.6 DETECTION LIMIT (DL) -The minimum concentration of an analyte that can be identified,
measured and reported with 99% confidence that the analyte concentration is greater than
zero. This is a statistical determination (Sect. 9.2.7), and accurate quantitation is not
expected at this level.2
3.7 LABORATORY DUPLICATES (LDs) - Two sample aliquots (LDi and LD2), from a single
field sample bottle, and analyzed separately with identical procedures. Analyses of LDi and
LD2 indicate precision associated specifically with laboratory procedures by removing
variation contributed from sample collection, preservation, and storage procedures.
3.8 LABORATORY FORTIFIED BLANK (LFB) - An aliquot of reagent water or other blank
matrix to which a known quantity of the method analyte is added. The LFB is analyzed
exactly like a sample, including the preservation procedures in Section 8.1. Its purpose is to
determine whether the methodology is in control, and whether the laboratory is capable of
making accurate and precise measurements.
3.9 LABORATORY FORTIFIED SAMPLE MATRIX (LFSM) - An aliquot of a field sample to
which a known quantity of the method analyte is added. The LFSM is processed and
analyzed exactly like a field sample, and its purpose is to determine whether the field sample
matrix contributes bias to the analytical results. The background concentration of the analyte
in the field sample matrix must be determined in a separate aliquot and the measured value in
the LFSM corrected for native concentrations.
3.10 LABORATORY FORTIFIED SAMPLE MATRIX DUPLICATE (LFSMD) - A second
aliquot of the field sample used to prepare the LFSM, which is fortified and analyzed
identically to the LFSM. The LFSMD is used instead of the Laboratory Duplicate to assess
method precision and accuracy when the occurrence of the target analyte is infrequent.
3.11 LABORATORY FORTIFIED SYNTHETIC SAMPLE MATRIX (LF S SM) - Aliquots of the
LSSM which are fortified with perchlorate (Sect. 7.2.2). These QC samples are used, during
an Analysis Batch, to confirm the integrity of the trapping efficiency of the concentrator
column and that the analyst has adequate resolution between the common anions and
perchlorate in high ionic matrices. The LFSSM samples are treated like the CCCs and are
not sterile filtered.
3.12 LABORATORY REAGENT BLANK (LRB) - An aliquot of reagent water or other blank
matrix that is treated exactly as a sample including exposure to all filtration equipment,
storage containers and internal standards. The LRB is used to determine if the method
analyte or other interferences are present in the laboratory environment, the reagents, or the
apparatus.
3.13 LABORATORY SYNTHETIC SAMPLE MATRIX (LSSM) - An aliquot of reagent water
that is fortified with 1000 mg/L of chloride, bicarbonate and sulfate. This solution is
representative of a drinking water containing 3000 mg/L of common anions.
314.1-4
-------
EPA815-R-05-009
3.14 LABORATORY SYNTHETIC SAMPLE MATRIX BLANK (LS 8MB) - An aliquot of the
LSSM that is processed like a field sample and is used to determine if the method analyte or
other interferences are present in the LSSMSS solution. It is also used to determine whether
the methodology is in control in terms of low system background.
NOTE: The LSSMB is processed through all sample collection steps outlined in Section
8.1. The LSSMB must be sterile filtered.
3.15 LABORATORY SYNTHETIC SAMPLE MATRIX FORTIFICATION SOLUTION
(LSSMFS) - A dilution of the LSSMSS is prepared to facilitate the addition of sodium to all
field samples in an accurate manner without necessitating volume correction (Sect. 7.2.3).
3.16 LABORATORY SYNTHETIC SAMPLE MATRIX STOCK SOLUTION (LSSMSS) - The
LSSMSS contains the common anions chloride, sulfate and bicarbonate at 25.0 g/L. This
solution is used in the preparation of all CAL and QC samples (Sect. 7.2.2).
3.17 LOWEST CONCENTRATION MINIMUM REPORTING LEVEL (LCMRL) - The single-
laboratory LCMRL is the lowest true concentration for which the future recovery is predicted
to fall between 50 and 150 percent recovery with 99% confidence. l
3.18 MATERIAL SAFETY DATA SHEET (MSDS) - Written information provided by vendors
concerning a chemical's toxicity, health hazards, physical properties, fire, and reactivity data
including storage, spill, and handling precautions.
3.19 MINIMUM REPORTING LEVEL (MRL) - The minimum concentration that can be
reported by a laboratory as a quantified value for the target analyte in a sample following
analysis. This defined concentration must meet the criteria defined in Section 9.2 and must
be no lower than the concentration of the lowest calibration standard for the target analyte.
3.20 PRIMARY DILUTION STANDARD SOLUTION (PDS) - A solution containing the
method analyte prepared in the laboratory from stock standard solutions and diluted as
needed to prepare calibration solutions and other analyte solutions.
3.21 QUALITY CONTROL SAMPLE (QCS) - A solution containing the method analyte at a
known concentration that is obtained from a source external to the laboratory and different
from the source of calibration standards. The QCS is used to verify the calibration
standards/curve integrity.
3.22 REAGENT WATER (RW) - Purified water which does not contain any measurable quantity
of the target analyte or interfering compounds at or above 1/3 the MRL.
3.23 SECONDARY DILUTION STANDARD SOLUTION (SDS) - A solution containing the
method analyte prepared in the laboratory from the PDS and diluted as needed to prepare
calibration solutions and other analyte solutions.
3.24 STOCK STANDARD SOLUTION (SSS) - A concentrated solution containing the method
analyte prepared in the laboratory using assayed reference materials or purchased from a
reputable commercial source.
314.1-5
-------
EPA815-R-05-009
4. INTERFERENCES
4.1 Interferences can be divided into three different categories: (i) direct chromatographic co-
elution, where an interfering analyte response is observed at very nearly the same retention
time (RT) as the target analyte; (ii) concentration dependant co-elution, which is observed
when the response of higher than typical concentrations of the neighboring peak overlaps into
the retention window of the target analyte; and (iii) ionic character displacement, where
retention times may significantly shift due to the influence of high ionic strength matrices
(high mineral content or Total Dissolve Solids) overloading the exchange sites on the column
and significantly shortening the target analyte's retention time.
4.1.1 A direct chromatographic co-elution may be solved by changing columns, eluent
strength, modifying the eluent with organic solvents (if compatible with 1C columns),
changing the detection systems, or selective removal of the interference with
pretreatment. Sample dilution will have little to no effect. The analyst must verify that
these changes do not induce any negative affects on method performance by repeating
and passing all the QC criteria as described in Section 9.2.
4.1.2 Sample dilution may resolve some of the difficulties if the interference is the result of
either concentration dependant co-elution or ionic character displacement, but it must be
clarified that sample dilution will alter your MRL by a proportion equivalent to that of
the dilution. Therefore, careful consideration of DQOs should be given prior to
performing such a dilution.
4.2 Method interferences may be caused by contaminants in solvents, reagents (including reagent
water), sample bottles and caps, and other sample processing hardware that lead to discrete
artifacts and /or elevated baselines in the chromatograms. All items such as these must be
routinely demonstrated to be free from interferences (less than l/j the perchlorate MRL) under
the conditions of the analysis by analyzing LRBs and LSSMBs as described in Section 9.2.1.
Subtracting blank values from sample results is not permitted.
4.3 Matrix interferences may be caused by contaminants that are present in the sample. The
extent of matrix interferences will vary considerably from source to source, depending upon
the nature of the water. Water samples high in organic carbon or TDS may have elevated
baselines or interfering peaks.
4.4 Equipment used for sample collection and storage has the potential to introduce interferences.
The potential for interferences from these devices must be investigated during the Initial
Demonstration of Capability (Sect. 9.2) by preparing and analyzing a LRB and LSSMB. This
procedure should be repeated each time that a new brand or lot of devices are used to ensure
that contamination does not hinder analyte identification and quantitation.
4.5 This method utilizes a confirmation column that has a separation mechanism that is
sufficiently different for the primary column so that perchlorate may be confirmed. The
suggested primary column, the lonPac AS 16, has a column chemistry that is based on a low
cross-link vinyl aromatic quaternary monomer. It was designed to provide good
chromatographic performance for polarizable inorganic anions such as perchlorate with
moderate concentration hydroxide eluents. Although less polarizable than inorganic species
314.1-6
-------
EPA815-R-05-009
such as perchlorate, such aromatic species show enhanced retention due to interaction with the
pi electrons of the aromatic backbone. The suggested confirmation column, the lonPac AS20,
has a column chemistry that is based on a cross-linked quaternary condensation polymer
completely free of any pi electron containing substituents. As such, it exhibits selectivity for
polarizable anions which is complementary to the AS 16, but because of the absence of any pi
electron character, retention of aromatic anionic species is greatly diminished relative to that
of the AS 16.
4.5.1 One component that has been shown by 1C and IC-MS to potentially co-elute with
perchlorate on the lonPac AS16 column when using EPA Method 314.0 protocols is 4-
chlorobenzenesulfonic acid (4-C1 BSA).3 As shown in Figure 1, with EPA Method
314.1 protocols, there is some resolution of the two components on the AS16 column and
the lonPac AS20 column provides excellent separation of perchlorate and 4-C1 BSA.
5. SAFETY
5.1 The toxicity or carcinogenicity of each reagent used in this method has not been precisely
defined. Each chemical should be treated as a potential health hazard, and exposure to these
chemicals should be minimized. Each laboratory is responsible for maintaining an awareness
of OSHA regulations regarding safe handling of chemicals used in this method. A reference
file of MSDSs should be made available to all personnel involved in the chemical analysis.
Additional references to laboratory safety are available.4"6
6. EQUIPMENT AND SUPPLIES (References to specific brands or catalog numbers are included
for illustration only and do not imply endorsement of the product.)
6.1 NON-STERILE SAMPLE CONTAINERS - 125-mL brown Nalgene bottles (Fisher Cat. No.
03-313-3C or equivalent).
6.2 STERILE SAMPLE CONTAINERS - 125-mL sterile high-density polyethylene (HOPE)
bottles (I-Chem 125-mL sterile HOPE bottle, Fisher Cat. No. N411-0125 or equivalent).
6.3 SAMPLE FILTERS - Sterile sample filters (Corning 26-mm surfactant free cellulose acetate
0.2-um filter, Fisher Cat. No. 09-754-13 or equivalent). If alternate filters are used they
should be certified as having passed a bacterial challenge test. 7 In addition, if alternate filters
or different lots of the recommended filters are used, they must be tested using a LSSMB and
a LFSSM fortified at the MRL as outlined in Section 9.2 to insure that they do not introduce
interferences or retain perchlorate.
6.4 SYRINGES - 20-mL sterile, disposable syringes (Henke Sass Wolf 20 mL Luer lock, Fisher
Cat. No. 14-817-33 or equivalent).
6.5 VOLUMETRIC FLASKS - Class A, suggested sizes include 10, 50, 100, 250, 500 and 1000
mL for preparation of standards and eluents.
6.6 GRADUATED CYLINDERS - Suggested sizes include 25 and 1000 mL.
6.7 AUTO PIPETTES - Capable of delivering variable volumes from 1.0 uL to 2500 uL.
314.1-7
-------
EPA815-R-05-009
6.8 ANALYTICAL BALANCE - Capable of weighing to the nearest 0.0001 g.
6.9 ION CHROMATOGRAPHY SYSTEM WITH SUPPRESSED CONDUCTIVITY
DETECTION (1C) - A Dionex model DX500 1C was used to collect the data presented in this
method. Alternative 1C systems can be used provided all the QC criteria listed in Section 9
are met. The 1C system must have a thermostatically controlled column heater and be capable
of operating above room temperature (35 °C) and include an ion chromatographic pump and
all required accessories including, analytical, concentrator and guard columns,
chromatography module, eluent generator, compressed gasses, autosampler, suppressor,
conductivity detector, and a computer-based data acquisition and control system.
Additionally, the system must be capable of performing inline sample pre-concentration and
matrix elimination steps.
6.9.1 CONCENTRATOR COLUMN - 1C column, 4.0 x 35-mm (Dionex Cryptand Cl or
equivalent). Any concentrator column that provides effective retention/trapping and
eventual release of perchlorate while providing the resolution, peak shape, capacity,
accuracy, and precision (Sect. 9.2) may be used. However, prior to use, the capacity of
the concentrator column must be evaluated as per Section 11.4.
6.9.2 PRIMARY GUARD COLUMN - 1C column, 2.0 x 50-mm (Dionex IonPac®AG16 or
equivalent). Any column that provides adequate resolution, peak shape, capacity,
accuracy, and precision (Sect. 9.2) may be used.
6.9.3 CONFIRMATION GUARD COLUMN - 1C column, 2.0 x 50-mm (Dionex
IonPac®AG20 or equivalent). Any column that provides adequate resolution, peak shape,
capacity, accuracy, and precision (Sect. 9.2) may be used. The separation mechanism for
the confirmation guard column must differ from the primary column.
6.9.4 PRIMARY ANALYTICAL COLUMN - 1C column, 2.0 x 250-mm (Dionex
IonPac®AS16 or equivalent). Any column that provides adequate resolution, peak shape,
capacity, accuracy, and precision (Sect. 9.2) may be used.
6.9.5 CONFIRMATION ANALYTICAL COLUMN - 1C column, 2.0 x 250-mm (Dionex
IonPac®AS20 or equivalent). Any column that provides adequate resolution, peak shape,
capacity, accuracy, and precision (Sect. 9.2) may be used. The separation mechanism for
the confirmation analytical column must differ from the primary column.
6.9.6 AUTOSAMPLER - A Dionex AS40 autosampler (or equivalent) is required to perform
the sample pre-concentration/matrix elimination steps. The method program must
include a timing sequence to allow the autosampler to load two sample vials before the
concentrator column is switched in-line to separate and detect perchlorate. The first
sample vial contains the sample (2.0 mL) and the second vial contains the rinse solution
(1.0 mL of 10 mM NaOH), with the filter cap raised to signify a rinse vial. The method
programs for the AS16 and AS20 columns are listed in Table 1 A. The method timing
sequence for the methods is listed in Table IB.
6.9.7 ELUENT GENERATOR - An eluent generator (Dionex EG50 or equivalent) with a
sodium cartridge (EluGen ® PN 058908 or equivalent) is used to prepare the sodium
314.1-8
-------
EPA815-R-05-009
hydroxide eluent for this method. An equivalent eluent generator may be used and/or
manually prepared eluents may also be used provided that adequate resolution, peak
shape, capacity, accuracy, and precision (Sect. 9.2) are obtained. Care must be exercised
with manually prepared sodium hydroxide eluents to prevent formation of carbonate in
the eluent from exposure to the atmosphere, which can dramatically alter the
chromatography.
NOTE: The Cryptand concentrator columns use manipulation of column capacity as part
of the mechanism for separation. The counter ion in the eluent has a strong influence on
both the concentrator column capacity and the capacity modification kinetics. In this
work, sodium is used to establish the optimal capacity of the Cryptand concentrator
column and consequently a sodium cartridge (Dionex PN 058908 or equivalent) rather
than a potassium cartridge MUST be used with the EG50.
6.9.8 ANION SUPPRESSOR DEVICE - The data presented in this method were generated
using a Dionex Ultra II Anion Self-Regenerating Suppressor (2-mm ASRS, PN 061562)
for electrolytic suppression of the eluent. Equivalent suppressor devices may be utilized
providing a comparable conductivity MRL and DL are achieved and adequate baseline
stability is attained as measured by a baseline noise of no more than 5 nS per minute over
the background conductivity.
NOTE: The conductivity suppressor was set to perform electrolytic suppression at a
current setting of 100 mA using the external water mode. Since unacceptable baseline
stability was observed on the conductivity detector using the Ultra II ASRS in recycle
mode, the external water mode must be used
6.9.9 CONDUCTIVITY DETECTOR - Conductivity cell (Dionex CD20 or equivalent)
capable of providing data as required in Section 9.2.
6.9.10 CHROMATOGRAPHY MODULE - A chromatography module (Dionex LC30 or
equivalent) capable of maintaining the columns, suppressor and conductivity cell at 35 °C
is required.
6.9.11 DATA SYSTEM - An interfaced data system such as Dionex, Chromeleon Version 6.0
(or equivalent) is required to acquire, store, and output conductivity data. The computer
software should have the capability of processing stored conductivity data by recognizing
a peak within a given retention time window. The software must allow integration of the
peak area of any specific peak between specified time limits. The software must be able
to construct a linear regression or quadratic calibration curve, and calculate analyte
concentrations.
7. REAGENTS AND STANDARDS
7.1 REAGENTS - Reagent grade or better chemicals should be used in all tests. Unless
otherwise indicated, it is intended that all reagents will conform to the specifications of the
Committee on Analytical Reagents of the American Chemical Society (ACS), where such
specifications are available. Other grades may be used, provided it is first determined that
314.1-9
-------
EPA815-R-05-009
the reagent is of sufficiently high purity to permit its use without lessening the quality of the
determination.
7.1.1 REAGENT WATER (RW) - Purified water which does not contain any measurable
quantity of the target analyte or interfering compounds at or above 1/3 the perchlorate
MRL. The purity of the water required for this method cannot be overly emphasized.
The reagent water used during method development was generated from tap water using
a Millipore ELIX-3 followed by a Millipore Gradient A10 system. The water should
contain no particles larger than 0.20 microns.
7.1.2 ELUENT SOLUTION - Sodium hydroxide eluent concentrations of 0.50, 65 and 100
mM are automatically prepared using the EG50 eluent generator and/or manually
prepared (Sect. 6.9.7).
7.1.3 SODIUM BICARBONATE - (NaHCO3, CASRN 497-19-8) - Fluka Cat. No. 71627 or
equivalent.
7.1.4 SODIUM CHLORIDE - (NaCl, CASRN 7647-14-5) - Fisher Cat. No. S-271 or
equivalent.
7.1.5 SODIUM SULFATE - (Na2SO4, CASRN 7757-82-6) -Fluka Cat. No. 71959 or
equivalent.
7.2 STANDARD SOLUTIONS - When a compound purity is assayed to be 96 percent or greater,
the weight can be used without correction to calculate the concentration of the stock standard.
Solution concentrations listed in this section were used to develop this method and are
included as an example. Even though stability times for standard solutions are suggested
in the following sections, laboratories should use standard QC practices to determine
when their standards need to be replaced.
7.2.1 PERCHLORATE STANDARD SOLUTIONS - Obtain the analyte as a solid standard of
NaClO4 or as a commercially prepared standard from a reputable standard manufacturer.
Prepare the Perchlorate Stock and Dilution Solutions as described below.
7.2.1.1 PERCHLORATE STOCK STANDARD SOLUTION (SSS) (1000 mg/L C1O4") - To
prepare this solution from a solid NaClO4 standard, weigh out 123.1 mg of NaClO4
into a 100-mL volumetric flask and dilute to volume with reagent water. When
stored in opaque, plastic storage bottles, the resulting stock solution may be stable for
up to one year.
7.2.1.2 PERCHLORATE PRIMARY DILUTION SOLUTION (PDS) (10.0 mg/L C1O4") -
Prepare the Perchlorate PDS by adding 1.00 mL of the Perchlorate SSS to a 100-mL
volumetric flask and dilute to volume with reagent water. This solution is used to
prepare the Secondary Dilution Solution, the Perchlorate Fortification Solutions and
the Calibration Solutions below. When stored in opaque, plastic storage bottles, the
resulting solution is stable for at least one month.
314.1-10
-------
EPA815-R-05-009
7.2.1.3 PERCHLORATE SECONDARY DILUTION SOLUTION (SDS) (1.00 mg/L C1O4')
- Prepare the 1.00 mg/L Perchlorate SDS by adding 10.0 mL of the Perchlorate PDS
to a 100-mL volumetric flask and dilute to volume with reagent water. This solution
is used to prepare the Perchlorate Fortification Solutions, CAL and CCC Standards
listed below. When stored in opaque, plastic storage bottles, the resulting solution is
stable for at least one month.
7.2.1.4 PERCHLORATE FORTIFICATION SOLUTIONS (PFS) (50, 200 and 500 |ig/L) -
The Perchlorate Fortification Solutions are prepared by dilution of the Perchlorate
SDS and are used to fortify the Laboratory Fortified Blank (LFB), the Laboratory
Fortified Synthetic Sample Matrix (LFSSM), the Laboratory Fortified Sample Matrix
(LFSM) and the Laboratory Fortified Sample Matrix Duplicate (LFSMD) with
perchlorate. It is recommended that multiple concentrations be prepared so that the
fortification levels can be rotated or adjusted to the concentration of target analyte in
the native samples. When stored in opaque, plastic storage bottles, the resulting
solutions are stable for at least one month. A 20-uL aliquot of each PFS added to a
2.0-mL sample volume yield a perchlorate concentration of 0.50, 2.0 and 5.0 ug/L,
respectively.
.2 LABORATORY SYNTHETIC SAMPLE MATRIX STOCK SOLUTION (LSSMSS) -
Prepare a LSSMSS that contains the common anions chloride, sulfate and bicarbonate at
25.0 g/L as follows. This solution is used in the preparation of all QC samples. A
dilution of the LSSMSS is used to fortify all samples (Sect. 7.2.3).
7.2.2.1 Weigh out 3.44 g of NaHCO3, 3.72 g of Na2SO4, and 4.00 g of NaCl (Fluka 1627,
Fluka 71959, Fisher S-271, respectively or equivalent). Quantitatively transfer these
to a 100-mL volumetric flask and dilute to volume using reagent water. This solution
is used to add 100 mg/L of the LSSM to all blanks, CALs and CCCs and all field
samples.
NOTE: EPA Method 314.0 incorporated a synthetic sample matrix containing 1000
mg/L of chloride, carbonate and sulfate that yielded a pH of approximately 10.
Method 314.1 uses bicarbonate which yields a pH of approximately 8.6, which more
closely resembles a finished drinking water. It should be noted that pH 10 carbonate
is a stronger eluent that could cause break-through of perchlorate on the Cryptand
concentrator column under conditions listed for this method and should therefore not
be used to prepare this solution.
.3 LABORATORY SYNTHETIC SAMPLE MATRIX FORTIFICATION SOLUTION
(LSSMFS) - As noted in Sect. 11.4, the capacity of the Cryptand concentrator column is
set with sodium. A dilution of the LSSM Stock Solution is prepared to facilitate the
addition of sodium to all field samples in an accurate manner yet without necessitating
volume correction. Prepare an LSSMFS that contains the common anions chloride,
sulfate and bicarbonate at 12.5 g/L as follows.
Add 50.0 mL of the LSSMSS to a 100-mL volumetric flask and dilute to volume
using reagent water. The LSSMFS solution is used to add 100 mg/L of the common
anions to all field samples (17 |jL/2.0 mL of field sample).
314.1-11
-------
EPA815-R-05-009
7.2.4 CALIBRATION BLANK (CB) - Prepare a CB that contains 100 mg/L of the common
anions to ensure effective trapping of the perchlorate by adding 400 |jL of the LSSMSS
to 100 mL of RW as indicated in the Table below. The CB is used only during the
initial calibration to ensure that no perchlorate or interferences are present in the
CAL standards (containing 100 [tgfL of the common anions) prior to calibration.
The CB is not sterile filtered prior to analysis.
7.2.5 LABORATORY FORTIFIED BLANK (LFB) - Prepare an LFB that contains 100 mg/L
of the common anions by adding 400 [iL of the LSSMSS to 100 mL of RW and
fortifying the LFB with the appropriate volume of perchl orate PDS or SDS as indicated
in the Table below. The LFB must be sterile filtered prior to analysis.
7.2.6 LABORATORY SYNTHETIC SAMPLE MATRIX BLANK (LSSMB) - Prepare the
LSSMB by adding 4000 [iL of LSSMSS to 100 mL of RW as indicated in the Table
below. The LSSMB must be sterile filtered prior to analysis.
7.3 CALIBRATION STANDARDS (CAL) - Prepare a calibration curve from dilutions of the
Perchlorate PDS, the Perchlorate SDS, and the LSSMSS using a minimum of five
Calibration Standards, which span the concentration range of interest. The lowest CAL
standard must be at or below the MRL. An example of the dilutions used to prepare the CAL
standards used to collect the data in Section 17, are shown in the Table below.
NOTE: CAL standards are not processed with the sample collection devices or protocols.
This step must be omitted for the CALs in order to identify any potential losses associated
with the sample filtration or collection protocols.
7.4 CONTINUING CALIBRATION CHECK STANDARDS (CCC) - Prepare the CCC
standards from dilutions of the Perchlorate PDS, the Perchlorate SDS, and the LSSMSS. An
example of the dilutions used to prepare the CCCs that were used to collect the data in Section
17 are shown in the Table below.
NOTE: CCC standards are not processed with the sample collection devices or protocols.
This step must be omitted for the CCCs in order to identify any potential losses associated
with the sample filtration or collection protocols.
7.5 LABORATORY FORTIFIED SYNTHETIC SAMPLE MATRIX CCC STANDARDS -- In
order to continually monitor the integrity of the trapping efficiency of the concentrator
column throughout an Analysis Batch, the CCCs are also prepared in a 1000 mg/L common
anion synthetic matrix. These solutions are termed Laboratory Fortified Synthetic Sample
Matrix (LFSSM) CCCs and are analyzed following the normal CCCs during the Analysis
Batch. An example of the dilutions used to prepare the LFSSM CCCs that were used to
collect the data in Section 17, are shown in the Table below. LFSSM CCCs are processed
through all sample collection devices and protocols.
314.1-12
-------
EPA815-R-05-009
PREPARATION OF CAL, CCC AND LFSSM CCC STANDARDS
CALand CCC
Levels
CB
LSSMB
CAL 1
CAL 2
CAL 3
CAL 4
CAL 5
CAL 6
Low-CCC
Mid-CCC
High-CCC
Low-LFSSM CCC
Mid-LFSSM CCC
High-LFSSM CCC
Vol. of
C1O4
PDS
(uL)
30
50
100
50
100
50
100
Vol. of
C1O4
SDS
(uL)
30
50
100
50
50
Vol. of
LSSMSS
(uL)
400
4000
400
400
400
400
400
400
400
400
400
4000
4000
4000
Final
Vol. of
Std.
(mL)
100
100
100
100
100
100
100
100
100
100
100
100
100
100
Final
Cone, of
Common
Anions
(mg/L)
100
1000
100
100
100
100
100
100
100
100
100
1000
1000
1000
Final
Cone, of
C1O4
(Hg/L)
0.0
0.0
0.30
0.50
1.00
3.00
5.00
10.0
0.50
5.0
10
0.50
5.0
10
8. SAMPLE COLLECTION, PRESERVATION, AND STORAGE
8.1 SAMPLE COLLECTION
8.1.1 Grab samples must be collected in accordance with conventional sampling practices.8
8.1.2 When sampling from a cold water tap, open the tap and allow the system to flush until the
water temperature has stabilized (usually approximately 3 to 5 minutes). Collect a
representative sample from the flowing system using a beaker of appropriate size. Use
this bulk sample to generate individual samples as needed. A volume of at least 20-mL is
required for each individual sample.
8.1.3 When sampling from an open body of water, fill a beaker with water sampled from a
representative area. Use this bulk sample to generate individual samples as needed. A
volume of at least 20-mL of filtered sample is required for each individual sample.
8.1.4 Once representative samples are obtained (at the time of collection), they must be sterile
filtered (Sect. 8.1.4.1) to remove any native microorganisms. Perchlorate is known to be
susceptible to microbiological degradation by anaerobic bacteria.9 Samples are sterile
filtered to remove microbes and stored with headspace to reduce the potential for
degradation by any remaining anaerobic organisms.
8.1.4.1 Remove a sterile syringe (Sect. 6.4) from its package and draw up approximately 25
mL of the bulk sample (fill the syringe). Remove a sterile syringe filter (Sect 6.3)
from its package without touching the exit Luer connection. Connect the filter to the
syringe making sure that no water from the syringe drops on the exterior of the filter.
314.1-13
-------
EPA815-R-05-009
Depress the syringe plunger gently and discard the first 3-5 mL. Open a sterile
sample container (Sect. 6.2) without touching the interior. Using gentle pressure,
pass the sample through the filter into the sample container. During this process do
not let the syringe or filter make contact with the sample container. Following
filtration, seal the sample container tightly, label and prepare the container for
shipment. Syringes and filters are single use items and must be discarded after each
sample.
8.2 SAMPLE SHIPMENT AND STORAGE - Field samples must be chilled during shipment and
must not exceed 10 °C during the first 48 hours after collection. Field samples should be
confirmed to be at or below 10 °C when they are received at the laboratory. Field samples
stored in the lab must be held at or below 6 °C until analysis, but should not be frozen.
8.3 SAMPLE HOLDING TIMES - Field samples that are collected and stored as described in
Sections 8.1 and 8.2 may be held for 28 days.
9. QUALITY CONTROL
9.1. Quality Control requirements include the Initial Demonstration of Capability (IDC) and
ongoing QC requirements that must be met when preparing and analyzing field samples. This
section describes each QC parameter, their required frequency, and the performance criteria
that must be met in order to meet EPA data quality objectives. The QC criteria discussed in
the following sections are summarized in Section 17, Tables 5 and 6. These QC requirements
are considered the minimum acceptable QC criteria. Laboratories are encouraged to institute
additional QC practices to meet their specific needs.
9.1.1 METHOD MODIFICATIONS - The analyst is permitted to modify the 1C system,
columns and separation conditions (Sect. 6.9). However, each time such method
modifications are made, the analyst must repeat the procedures of the IDC (Sect. 9.2). In
addition, if an alternate concentrator column is used, the procedure outlined in Section
11.4 MUST be completed before the IDC is initiated.
9.2 INITIAL DEMONSTRATION OF CAPABILITY (IDC) - The IDC must be successfully
performed prior to analyzing any field samples. Prior to conducting the IDC, the analyst must
first generate an acceptable Initial Calibration following the procedure outlined in Section
10.2. Requirements for the IDC are described in the following sections and are summarized
in Table 5.
9.2.1 DEMONSTRATION OF LOW SYSTEM BACKGROUND - Analyze a Laboratory
Synthetic Sample Matrix Blank (LSSMB) processed through all sample collection steps
outlined in Section 8.1. The LSSMB must be sterile filtered. Confirm that the LSSMB
is reasonably free of contamination and that the criteria in Section 9.3.1 and 9.3.2 are
met.
NOTE: It is Good Laboratory Practice to include a blank in the calibration of any
instrument. As well, the method should be checked for carry-over by analyzing a LSSMB
blank immediately following the highest CAL standard. If this LSSMB sample does not
314.1-14
-------
EPA815-R-05-009
meet the criteria outlined in Section 9.3.1 then carry-over is present and should be
identified and eliminated.
9.2.2 DEMONSTRATION OF PRECISION - Prepare and analyze 7 replicate LFBs and
LFSSMs fortified near the midrange of the initial calibration curve. All samples must be
fortified and processed using the sample collection protocols described in Section 8.1.
The percent relative standard deviation (%RSD) of the results of the replicate analyses
must be < 20 percent.
n. _, nT^ Standard Deviation of Measured Concentrations ,
% RSD = x 100
Average Concentration
9.2.3 DEMONSTRATION OF ACCURACY - Using the same set of replicate data generated
for Section 9.2.2, calculate average recovery. The average recovery of the replicate
values must be within ± 25 percent of the true value.
n. _. Average Measured Concentration ,nn
% Recovery = x 100
Fortified Concentration
9.2.4 MINIMUM REPORTING LEVEL (MRL) CONFIRMATION - Establish a target
concentration for the MRL based on the intended use of the method. Establish an initial
calibration following the procedure outlined in Section 10.2. The lowest calibration
standard used to establish the initial calibration (as well as the low-level CCC) must be at
or below the concentration of the MRL. Establishing the MRL concentration too low
may cause repeated failure of ongoing QC requirements. Confirm or validate the MRL
following the procedure outlined below.
9.2.4.1 Fortify and analyze seven replicate Laboratory Fortified Blanks at the proposed MRL
concentration. All samples must be fortified and processed using the sample
collection protocols described in Section 8.1. Calculate the mean (Mean) and
standard deviation (S) for these replicates. Determine the Half Range for the
prediction interval of results (HRPIR) using the equation below.
HRPIR = 3.963S
where S is the standard deviation, and 3.963 is a constant value for seven replicates.
9.2.4.2 Confirm that the upper and lower limits for the Prediction Interval of Result (PIR =
Mean j_ HRpiR) meet the upper and lower recovery limits as shown below.
The Upper PIR Limit must be < 150 percent recovery.
HRP1R ,
FortifiedConcentration
The Lower PIR Limit must be > 50 percent recovery.
314.1-15
-------
EPA815-R-05-009
Mean - HRPIK
PIR xlOO>50%
FortifiedConcentration
9.2.4.3 The MRL is validated if both the Upper and Lower PIR Limits meet the criteria
described above (Sect. 9.2.4.2). If these criteria are not met, the MRL has been set
too low and must be determined again at a higher concentration.
9.2.5 MRL CONFIRMATION IN THE 1000 mg/L LFSSM - Fortify and analyze seven
replicate LFSSMs fortified at the proposed MRL concentration. All samples must be
fortified and processed using the sample collection protocols described in Section 8.1.
Follow the steps outlined in Sections 9.2.4.1 to validate the MRL in the LFSSM. If these
criteria are not met, the MRL has been set too low and must be determined again at a
higher concentration.
9.2.6 CALIBRATION CONFIRMATION - Analyze a Quality Control Sample as described in
Section 9.4.1 to confirm the accuracy of the calibration standards/calibration curve.
9.2.7 DETECTION LIMIT DETERMINATION (optional) - While DL determination is not a
specific requirement of this method, it may be required by various regulatory bodies
associated with compliance monitoring. It is the responsibility of the laboratory to
determine ifDL determination is required based upon the DQOs.
Analyses for this procedure should be done over at least 3 days. Prepare at least 7
replicate fortified LFBs using the sample collection protocols described in Section 8.1.
Use the solutions described in Section 7.2.1.4 to fortify at a concentration estimated to be
near the DL. This fortification concentration may be estimated by selecting a
concentration at 2-5 times the noise level. The DLs in Table 2 were calculated from
LFBs fortified at 0.10 //g/L. Analyze the seven replicates through all steps of Section 11.
NOTE: If an MRL confirmation data set meets these requirements, a DL may be
calculated from the MRL confirmation data, and no additional analyses are necessary.
Calculate the DL using the following equation:
DL = St(n- 1, 1 - alpha = 0.99)
where:
t(n-1,1 - alpha = 0.99) = Student's t value for the 99% confidence level with n-1
degrees of freedom
n = number of replicates
S = standard deviation of replicate analyses.
NOTE: Do not subtract blank values when performing DL calculations.
9.3 ONGOING QC REQUIREMENTS - This section describes the ongoing QC criteria that
must be followed when processing and analyzing field samples. Table 6 summarizes these
requirements.
314.1-16
-------
EPA815-R-05-009
9.3.1 LABORATORY REAGENT BLANK (LRB) - A LRB (Sect. 3.12) is analyzed during
the IDC to confirm that potential background contaminants are not interfering with the
identification or quantitation of perchlorate. If the LRB produces a peak within the
retention time window of perchl orate that would prevent the determination of perchl orate,
identify the source of contamination and eliminate the interference before processing
field samples. Background from the method analyte or other contaminants that interfere
with the measurement of perchl orate must be below l/j of the MRL.
9.3.2 LABORATORY SYNTHETIC SAMPLE MATRIX BLANK (LSSMB) - A LSSMB
(Sect. 3.14) is required with each Analysis Batch and is used to confirm that potential
background contaminants are not in the LFSSM fortification solution and are not
interfering with the identification or quantitation of perchl orate. If the LSSMB produces
a peak within the retention time window for perchlorate that would prevent the
determination of perchlorate, determine the source of contamination and eliminate the
interference before processing field samples. The LSSMB must contain the LSSM at the
1000 mg/L concentration and must be sterile filtered. Background contamination must
be reduced to an acceptable level before proceeding. Background from the method
analyte or other contaminants that interfere with the measurement of perchlorate must be
below l/j of the MRL. Blank contamination may be estimated by extrapolation if the
concentration is below the lowest calibration standard. This procedure is not allowed for
field sample results as it may not meet the DQOs. If perchlorate is detected in the
LSSMB at concentrations equal to or greater than this level, then all data for perchlorate
must be considered invalid for all field samples in the Analysis Batch.
9.3.3 CONTINUING CALIBRATION CHECK STANDARDS (CCC) - CCC standards are
analyzed at the beginning of each Analysis Batch, after every ten field samples, and at the
end of the Analysis Batch. See Section 10.3 and Table 6 for concentration requirements
and acceptance criteria.
9.3.4 LABORATORY FORTIFIED SYNTHETIC SAMPLE MATRIX CCC STANDARDS -
CCC standards are prepared in the LFSSM at the same concentration as the CCC
Standards and analyzed at the same frequency as the CCCs. The LFSSM CCCs are used
to ensure the integrity of the sample pre-concentration/matrix elimination step and the
chromatographic separation of perchlorate from other interfering anionic species in very
high ionic matrices. See Section 10.3 and Table 6 for concentration requirements and
acceptance criteria.
9.3.5 LABORATORY FORTIFIED BLANK - The LFB is only required during the IDC (Sect.
9.2) and is not required to be included in the Analysis Batch due to the requirement for a
LSSMB to be analyzed at the start of each Analysis Batch (Sect. 9.3.2).
9.3.6 LABORATORY FORTIFIED SAMPLE MATRIX (LFSM) - An aliquot of a field
sample to which a known quantity of the method analyte is added. The LFSM is
processed and analyzed exactly like a sample, and its purpose is to determine whether the
sample matrix contributes bias to the analytical results. The background concentration of
the analyte in the sample matrix must be determined in a separate aliquot and the
measured value in the LFSM corrected for background concentrations.
314.1-17
-------
EPA815-R-05-009
9.3.6.1 Within each Analysis Batch, a minimum of one field sample is fortified as an LFSM
for every 20 field samples analyzed. The LFSM is prepared by spiking a field sample
with an appropriate amount of the Perchlorate Fortification Solution (Sect. 7.2.1.4).
The fortification should be delivered in the smallest volume possible to minimize
dilution of the sample. Select a spiking concentration that is equal to or greater than
the native concentration, if known. Use historical data and rotate through the
designated concentrations when selecting a fortifying concentration.
9.3.6.2 Calculate the percent recovery (%R) for the analyte using the equation.
%R= " x 100
A = measured concentration in the fortified field sample
B = measured concentration in the unfortified field sample
C = fortification concentration.
NOTE: If the fortified concentration is below the native concentration, the fortified
value is not considered valid. The reported value should be flagged to show that the
fortification level was lower than native concentration. However, the fortification
frequency requirement for the method will have been met and the analysis batch
data considered acceptable.
9.3.6.3 For field samples fortified at or above their native concentration, recoveries should
range between 75 and 125 percent, except for low-level fortifications less than or
equal to the MRL where 50 to 150 percent recoveries are acceptable. If the accuracy
of perchlorate falls outside the designated range, and the laboratory performance for
the analyte is shown to be in control in the CCCs and LFSSM CCCs, the recovery is
judged to be matrix biased. The result for the analyte in the unfortified field sample
is labeled suspect/matrix to inform the data user that the results are suspect due to
matrix effects.
9.3.6.3.1 Field samples that have an observed positive native perchlorate concentration
less than the MRL and are fortified at concentrations at or near the MRL
should be corrected for the native levels in order to obtain meaningful percent
recovery values. This is the only permitted use of analyte results below the
MRL.
9.3.7 LABORATORY DUPLICATE OR LABORATORY FORTIFIED SAMPLE MATRIX
DUPLICATE (LD or LFSMD) - Within each Analysis Batch, a minimum of one
Laboratory Duplicate (LD) or Laboratory Fortified Sample Matrix Duplicate (LFSMD)
must be analyzed. Laboratory Duplicates check the precision associated with laboratory
procedures. If target analytes are not routinely observed in field samples, a LFSMD
should be analyzed rather than a LD. LFSMDs check the precision associated with
laboratory procedures.
314.1-18
-------
EPA815-R-05-009
9.3.7.1 Calculate the relative percent difference (RPD) for duplicate measurements (LDi and
LD2) using the equation.
LD.-LD,
\ xlOO
(LD1+LD2)/2
9.3.7.2 RPDs for Laboratory Duplicates should be < 25 percent. Greater variability may be
observed when Laboratory Duplicates have analyte concentrations that are within a
factor of 2 of the MRL. At these concentrations Laboratory Duplicates should have
RPDs that are < 50 percent. If the RPD of any analyte falls outside the designated
range, and the laboratory performance for that analyte is shown to be in control in the
CCCs and LFSSM CCCs, the recovery is judged to be matrix influenced. The result
for that analyte in the unfortified field sample is labeled suspect/matrix to inform the
data user that the results are suspect due to matrix effects.
9.3.7.3 If a LFSMD is analyzed instead of a Laboratory Duplicate, calculate the relative
percent difference (RPD) for duplicate LFSMs (LFSM and LFSMD) using the
equation.
LFSM - LFSMD
xlOO
(LFSM + LFSMD)/2
93.7.4 RPDs for duplicate LFSMs should be < 25 percent. Greater variability may be
observed when LFSMs are fortified at analyte concentrations that are within a factor
of 2 of the MRL. LFSMs fortified at these concentrations should have RPDs that are
< 50 percent. If the RPD of any analyte falls outside the designated range, and the
laboratory performance for that analyte is shown to be in control in the CCCs and
LFSSM CCCs, the recovery is judged to be matrix influenced. The result for that
analyte in the unfortified field sample is labeled suspect/matrix to inform the data
user that the results are suspect due to matrix effects.
9.4 QUARTERLY QC REQUIREMENTS
9.4.1 QUALITY CONTROL SAMPLES (QCS) - As part of the IDC (Sect. 9.2), each time a
new Analyte PDS (Sect. 7.2.1.2) is prepared, every time the instrument is calibrated and
at least quarterly, analyze a QCS sample fortified near the midpoint of the calibration
range. The QCS sample should be from a source different than the source of the
calibration standards. If a second vendor is not available, then a different lot of the
standard should be used. The QCS should be prepared and analyzed just like a CCC.
Acceptance criteria for the QCS is identical to the mid- and high-level CCCs; the
calculated amount for the analyte must be + 25 percent of the true value. If measured
analyte concentrations are not of acceptable accuracy, check the entire analytical
procedure to locate and correct the problem.
314.1-19
-------
EPA815-R-05-009
10. CALIBRATION AND STANDARDIZATION
10.1 Demonstration and documentation of acceptable initial calibration for perchlorate is required
before any field samples are analyzed. If alternative instrumentation and/or concentrator
columns to those listed in this method are used, the procedure outlined in Section 11.4
MUST be completed before the calibration can be initiated. If the initial calibration is
successful, continuing calibration check standards are required at the beginning and end of
each Analysis Batch, as well as after every tenth field sample.
NOTE: CAL solutions and CCC standards are not processed with the sample
collection protocols. This step must be omitted for the CALs and CCCs in order to identify
any potential losses associated with the sample filtration or collection devices.
10.2 INITIAL CALIBRATION - Initial calibration is established during the IDC and may be
reestablished prior to analyzing field samples. However, it is permissible to verify the
calibration with daily CCCs and LFSSM CCCs. Calibration must be performed using peak
areas and the external standard technique. Calibration using peak heights is not permitted.
NOTE: In this method, the CB, LFB, CAL, QCS and CCC standards are prepared in RW
fortified with 100 mg/L of the LSSM to ensure optimal trapping of perchl orate. The CB,
LFB, LRB, CAL, QCS and CCC standards are not sterile filtered. The CB is used only in the
IDC. On the other hand, the LSSMB represents a drinking water matrix containing 3000
mg/L of common anions and is used in all Analysis Batches and must be sterile filtered.
10.2.1 INSTRUMENT CONDITIONS - Establish proper operating conditions. Operating
conditions used during method development are described in Section 17 Table 1A.
Conditions different from those described may be used if the IDC QC criteria in Section
9.2 are met.
10.3 CALIBRATION STANDARDS - Prepare a set of at least five CAL standards as described
in Section 7.3. The lowest concentration CAL standard must be at or below the MRL. The
MRL must be confirmed using the procedure outlined in Section 9.2.4, after establishing the
initial calibration.
10.3.1 CALIBRATION - The conductivity detector is calibrated using the external standard
technique. Calibration curves may be generated using the 1C data system through the use
of a first (linear) or second (quadratic) order calibration curves.
10.3.2 CALIBRATION ACCEPTANCE CRITERIA - The validation of the calibration is
determined by calculating the concentration of the analyte from each of the analyses used
to generate the calibration curve. Each calibration point, except the lowest (< MRL), for
the analyte should calculate to be 75 to 125 percent of its true value. The lowest point
should calculate to be 50 to 150 percent of its true value. If these criteria cannot be met,
the analyst will have difficulty meeting ongoing QC criteria. Corrective action must be
taken to reanalyze the calibration standards, restrict the range of calibration, or select an
alternate method of calibration.
314.1-20
-------
EPA815-R-05-009
10.4 CONTINUING CALIBRATION CHECK (CCC) STANDARDS - The CCCs verify the
calibration at the beginning and end of each group of analyses, as well as after every 10th
field sample during analyses. The LRBs, LFBs, LFSSMs, LFSMs, LFSMDs, CCCs and
LFSSM CCCs are not counted as field samples. The beginning CCCs for each Analysis
Batch must be at or below the MRL in order to verify instrument sensitivity and the accuracy
of the calibration curve prior to the analysis of any field samples. Subsequent CCCs should
alternate between a medium and high concentration.
NOTE: The analyst may chose to also run a mid-level CCC at the start of an Analysis Batch.
10.4.1 Inject an aliquot of the CCC standards and analyze with the same conditions used during
the initial calibration.
10.4.2 Calculate the concentration of the analyte in the CCC standards. The calculated amount
for the analyte for medium and high level CCCs must be ± 25 percent of the true value.
The calculated amount for the lowest CCC level for the analyte must be within ± 50
percent of the true value. If these conditions do not exist, then all data for the analyte
must be considered invalid, and remedial action (Sect. 10.4.4) should be taken which may
require recalibration. Any field samples that have been analyzed since the last acceptable
calibration verification and are still within holding time should be reanalyzed after
adequate calibration has been restored.
10.4.3 LABORATORY FORTIFIED SYNTHETIC SAMPLE MATRIX CCC STANDARDS -
As noted in Section 9.3.4, LFSSM CCCs Standards are prepared and analyzed to verify
the integrity of the concentrator column during the Analysis Batch to ensure that high
ionic strength drinking water matrices will not exceed the capacity of the concentrator
column. These QC samples are fortified at the same level and run at the same frequency
as the CCC Standards and are required to meet the same recovery criteria (Sect. 10.4.2).
10.4.4 REMEDIAL ACTION - Failure to meet CCC or LFSSM CCC QC performance criteria
may require remedial action. Maintenance such as confirming the integrity of the
trapping efficiency of the concentrator column and matrix elimination step or
regenerating or replacing the 1C columns will require re-calibration (Sect. 10.2).
11. PROCEDURE
11.1 Important aspects of this analytical procedure include proper field sample collection and
storage (Sect. 8.1), ensuring that the instrument is properly calibrated (Sect. 10.2) and that all
required QC are met (Sect. 9) during the Analysis Batch. This section describes the
procedures for field sample preparation and analysis. If alternative instrumentation and/or
concentrator columns to those listed in this method are used, the procedure outlined in
Section 11.4 MUST be followed prior to analyzing field samples.
11.2 SAMPLE PREPARATION
11.2.1 Collect and store field samples as described in Section 8.1.
314.1-21
-------
EPA815-R-05-009
11.2.2 Process all LSSMBs, LFSMs and LFSMDs using the sample collection protocols is
Section 8.1.
11.2.3 Transfer a 2.0-mL aliquot of each field or QC Sample to an autosampler vial. Add 17 uL
of the LSSMFS (Sect. 7.2.3) to all field sample autosampler vials. Place the autosampler
vial in the appropriate position.
11.2.4 For each QC standard and field sample to be analyzed, prepare a second autosampler vial
containing 1.0 mL of the lOmM NaOH rinse solution with the filter cap raised to signify
a rinse vial. Place the rinse vials in the autosampler rack, after every QC standard and
field sample.
11.3 SAMPLE ANALYSIS
11.3.1 Establish the instrument operating conditions as described in Table 1A of Section 17.
Confirm that the analyte retention times for the calibration standards are stable.
NOTE: The ionic strength of the common anion solution used to prepare the LFSSM
CCCs will cause these solutions to have shorter retention times (see Sect. 11.3.4.1).
11.3.2 Establish a valid initial calibration following the procedures outlined in Section 10.2 or
confirm that the calibration is still valid by running a low-level CCC as described in
Section 10.4. If establishing an initial calibration for the first time, complete the IDC as
described in Section 9.2.
11.3.3 Analyze field and QC samples at their required frequencies using the same conditions
used to collect the initial calibration. Table 7 shows an acceptable analytical sequence
that contains all method-required QC samples.
11.3.4 COMPOUND IDENTIFICATION - Establish an appropriate retention time window for
perchlorate to identify it in QC and field sample chromatograms.
11.3.4.1 High ionic strength matrices have the potential to cause an increase in background
conductivity and severe tailing as other anions elute from the column and cause
the perchlorate retention time to decrease.
NOTE: As a result of the difference in ionic strength of the 100 and 1000 mg/L
common anion matrices, the retention time for perchlorate in the 1000 mg/L
matrix is approximately 0.2 minutes shorter than in the 100 mg/L matrix (the
higher ionic strength matrix may act as a stronger eluent) using the conditions
outlined in Table 1 A. Since the ionic strength of drinking water matrices may
vary considerably, the RT window for perchlorate must be set wide enough to
account for the variability in the ionic strength of the drinking water
matrices and yet exclude any potential interfering peaks. A window of
approximately 0.4 minutes has been found to be acceptable; however setting the
window too wide may require additional analyses on the confirmation column.
314.1-22
-------
EPA815-R-05-009
11.3.4.2 COMPOUND CONFIRMATION - Field samples that have a perchlorate result
on the primary column at or above the MRL require confirmation with a second
analytical column that has a dissimilar separation mechanism. EPA Methods
331.0 and 332.0 can be used for confirmation of perchlorate results obtained using
EPA Method 314.1.
11.3.5 EXCEEDING CALIBRATION RANGE - The analyst must not extrapolate above the
established calibration range. If an analyte result exceeds the range of the initial
calibration curve, the field sample may be diluted with reagent water and the diluted field
sample re-injected (the LSSMFS must be added to the diluted field sample prior to
analysis). Incorporate the dilution factor into final concentration calculations. The
dilution will also affect the perchlorate MRL.
11.4 CONCENTRATOR COLUMN EVALUATION - This method was developed with a
Dionex Cryptand Cl concentrator column. Alternate columns are allowed, but prior to their
use, they must be evaluated to optimize sample injection volume, to confirm that the matrix
elimination step does not remove perchlorate, and to confirm that the perchlorate is
quantitatively transferred to and refocused on the guard column prior to separation on the
analytical column. The entire success of this method is totally dependent upon development
of column combinations that accomplish the aforementioned protocols. The procedure is
challenging and requires very experienced 1C chemists to evaluate alternative concentrator
column/guard and analytical column combinations using the procedures described below.
11.4.1 CONCENTRATOR COLUMN CAPACITY DETERMINATION - Any concentrator
column that provides effective retention/trapping and eventual release of perchlorate
while providing the resolution, peak shape, capacity, accuracy, and precision (Sect. 9.2)
may be used. However, prior to use, the capacity of the concentrator column must be
evaluated. The analyst must demonstrate the ability to load (or concentrate) at least 2.0
mL of a 5.0 |J,g/L perchlorate standard in the 1000 mg/L LFSSM (the loading volume
required to obtain the data presented in this method) without exceeding more than 80% of
the capacity of the concentrator column. This requirement ensures that the addition of
the 1000 mg/L of common anions to the field samples will not exceed the capacity of the
concentrator column.
11.4.1.1 Prepare 100-mL of the 5.0 |ig/L LFSSM CCC according to directions in Table
1. Load increasing volumes of the LFSSM CCC (1.0, 2.0, 3.0, 4.0 and 5.0-
mL, smaller increments may be used if desired) using the procedure outlined
for sample preparation and analysis sections (Sect. 11.2, 11.3). Observe when
perchlorate break-through occurs (i.e., no further increase in observed
perchlorate peak area or concentration). Plotting the peak area or
concentration versus load volume (as a histogram) will establish the volume at
which break-though of the perchlorate becomes evident. At this point, 100%
of the capacity of the concentrator column has been exceeded. Ensure that the
load volume to be used does not exceed the 80% restriction. It is
recommended that this procedure be reproduced at least twice to confirm the
break-through point.
314.1-23
-------
EPA815-R-05-009
11.4.2 EVALUATION OF MATRIX ELIMINATION CONDITIONS - Prior to use of a
concentrator column other than the one listed in this method, the matrix elimination
protocols must be evaluated in order to ensure that the perchlorate is retained on the
concentrator column while the interfering matrix anions are removed (to an acceptable
level) and sent to waste.
11.4.2.1 Once the load volume has been established, this can be accomplished by rinsing
the concentrated perchlorate (on the concentrator column) with different
concentrations and volumes of rinse solution. Prepare several weak NaOH rinse
solutions (0.50, 1.0 and 1.5 mM). Prepare several autosampler vials containing
the optimized volume of the 5.0 |ig/L LFSSM CCC. Prepare several autosampler
rinse vials containing different volumes of the NaOH rinse solutions (0.50, 1.0
and 1.5-mL) and analyze using the procedure outlined for sample preparation and
analysis sections (Sect. 11.2, 11.3). Choose a concentration and volume that will
meet the above criteria. The background conductivity must be less than 1.5 |j,S
when perchlorate elutes in order to obtain data similar to that reported in Tables 2,
3, and 4.
11.4.2.1.1 The rinse solution used to collect this data was 1.0 mL of 10 mM NaOH,
prepared from 50% NaOH by diluting 0.8 g of 50% NaOH to 1L with RW. In
order to prevent accumulation of carbonate in the rinse solution, the rinse
solution is stored, under helium, in a pressurized vessel fitted with a two-way
valve on the out line in order to withdraw the rinse solution as required. This
rinse solution is prepared fresh weekly.
11.4.3 EVALUATION OF WASH STEP CONDITIONS - Prior to use of a concentrator
column other than the one listed in this method, the wash step, which elutes the
perchlorate off the concentrator column and refocuses it at the head of the guard column,
must be evaluated. The wash step ensures quantitative transfer of the concentrated
perchlorate to the guard column head and minimizes band-broadening by ensuring that
the perchlorate is efficiently refocused on the guard column before the eluent strength is
increased to effect separation and detection of the perchlorate on the analytical column.
These steps are critical to method performance and were carefully optimized for the
Cryptand concentrator column during method development.
11.4.3.1 Once the load volume and rinse solution concentration and volume have been
established, evaluation of the wash step conditions is accomplished by removing
the guard and analytical columns from the system and connecting the concentrator
column directly to the conductivity detector. Use the EG50 to prepare the wash
solutions or use the manually prepared NaOH wash solutions (0.50, 1.0 and 1.5
mM). Using the optimized load volume and rinse solution determined above, use
the 5.0 |ig/L LFSSM CCC and rinse solution and modify the method to allow
different rinse times (10, 12 and 15 minutes) and concentrations of wash solution
to be evaluated using the procedure outlined for sample preparation and analysis
sections (Sect. 11.2, 11.3). Observe the time at which all the perchlorate has
eluted from the concentrator column (baseline returned minimum conductance).
The addition of a couple of minutes will ensure complete removal of perchlorate
from the concentrator column in all matrices. The 12 minute wash step for this
314.1-24
-------
EPA815-R-05-009
method provides a non-Gaussian peak that shows when the conductivity baseline
has returned to the minimum.
11.4.3.2 After establishing the optimal load volume, rinse solution concentration and
volume and wash time, concentration and volume, analyze a low-level CCC and
LFSSM CCC to ensure that the optimal conditions chosen provide acceptable
chromatography and peak shape and area for perchlorate.
11.4.3.2.1 The EG50 was used to prepare the wash solution (0.50 mM NaOH for 12
minutes) used to collect the data reported in Tables 2, 3, and 4. If the
manually prepared wash solution is used, the same precautions to prevent
accumulation of carbonate in the wash solution are required. This wash
solution is prepared fresh weekly.
12. DATA ANALYSIS AND CALCULATIONS
12.1 Identify the analyte present in the field and QC samples as described in Section 11.3.4.
NOTE: Since the ionic strength of the drinking water matrices can vary dramatically, and
perchlorate elutes on the trailing edge of the residual anions present in the sample, the
background conductivity is not the same for all injections. Consequently, the slope of the
baseline when perchlorate elutes may vary from sample to sample. As a result, it is quite
possible that the pre-set, auto- integration parameters may not start and stop peak integration
the same for every sample. Therefore, the analyst must thoroughly review all chromatograms
and some of the chromatograms may require manual integration of the perchlorate peak.
12.2 Calculate the perchlorate concentrations using the multi-point calibration established in
Section 10.2. Quantify only those values that fall between the MRL and the highest
calibration standard. Field samples with target analyte responses that exceed the highest
calibration standard require dilution and reanalysis (Sect. 11.3.5).
12.2.1 As noted in Section 9.3.2, it may be necessary to extrapolate below the MRL to estimate
contaminants in LRBs and LSSMBs and to correct for native levels of perchlorate below
the MRL when field samples are fortified at or near the MRL. These are the only
permitted use of analyte results below the MRL.
12.3 Calculations must utilize all available digits of precision, but final reported concentrations
should be rounded to an appropriate number of significant figures (one digit of uncertainty),
typically two, and not more than three significant figures.
12.4 Prior to reporting data, the laboratory is responsible for assuring that QC requirements have
been met or that any appropriate qualifier is documented.
13. METHOD PERFORMANCE
13.1 PRECISION, ACCURACY AND DETECTION LIMITS - Tables for these data are
presented in Section 17. Instrumental conditions are presented in Table 1A. The LCMRL for
perchlorate with both the AS 16 and AS20 columns is presented in Table 2 and was calculated
314.1-25
-------
EPA815-R-05-009
using a procedure described elsewhere.1 Single laboratory precision and accuracy data are
presented in Tables 3 and 4.
13.2 Figure 1 is a representative chromatogram showing the separation of perchlorate from 4-C1
BSA and Figure 2 shows a chromatogram of a surface and a ground water fortified with 1.0
Hg/L perchlorate and Figure 3 shows a chromatogram of 3.0 //g/L C1O4" in the 50, 500 and
lOOOmg/LLSSM.
14. POLLUTION PREVENTION
14.1 For information about pollution prevention that may be applicable to laboratory operations,
consult "Less is Better: Laboratory Chemical Management for Waste Reduction" available
from the American Chemical Society's Department of Government Relations and Science
Policy, 1155 16th Street N.W., Washington, D.C., 20036, or on-line at:
http: //www. up s. edu/community/storeroom/Chemi cal_Wastes/wastearti cl e s. htm.
15. WASTE MANAGEMENT
15.1 The analytical procedures described in this method generate relatively small amounts of waste
since only small amounts of reagents are used. The matrices of concern are finished drinking
water or source water. However, the Agency requires that laboratory waste management
practices be conducted consistent with all applicable rules and regulations, and that
laboratories protect the air, water, and land by minimizing and controlling all releases from
fume hoods and bench operations. Also, compliance is required with any sewage discharge
permits and regulations, particularly the hazardous waste identification rules and land disposal
restrictions. For further information on waste management, consult "The Waste Management
Manual for Laboratory Personnel" available from the American Chemical Society at the
address in Section 14.1, or on-line at: http://www.p2pays.org/ref/0l/text/00779/ch 15.htm.
16. REFERENCES
1. Revisions to the Unregulated Contaminant Monitoring Regulation for Public Water Systems,
Proposed Rule, 2004.
2. Glaser, J.A., D.L. Foerst, G.D. McKee, S.A. Quave, and W.L. Budde, "Trace Analyses for
Wastewaters", Environ. Sci. Technol.. 15 (1981) 1426_1435.
3. Personal Communication.
4. "Carcinogens - Working With Carcinogens," Department of Health, Education, and Welfare,
Public Health Service, Center for Disease Control, National Institute for Occupational Safety and
Health, Publication No. 77-206, Aug. 1977.
5. "OSHA Safety and Health Standards, General Industry," (29CFR1910), Occupational Safety and
Health Administration, OSHA 2206, (Revised, January 1976).
6. "Safety in Academic Chemistry Laboratories," American Chemical Society Publication,
Committee on Chemical Safety, 3rd Edition, 1979.
314.1-26
-------
EPA815-R-05-009
7. Blosse, P.T., Boulter, E.M., Sundaram, S., "Diminutive Bacteria Implications for Sterile
Filtration", Pall Corporation, East Hills, NY.
8. ASTM Annual Book of Standards, Part II, Volume 11.01, D3370-82, "Standard Practice for
Sampling Water," American Society for Testing and Materials, Philadelphia, PA, 1986.
9. Xu, J., Y. Song, B. Min, L. Steinberg, and B.E. Logan. 2003. Microbial degradation of
perchlorate: principles and applications. Environ. Engin. Sci, 20(5): 405-422.
314.1-27
-------
EPA815-R-05-009
17.0 TABLES, DIAGRAMS, FLOWCHARTS, AND VALIDATION DATA
TABLE LA. INSTRUMENTAL CONDITIONS
Standard Conditions and Equipment for Primary Analyses
Ion Chromatograph:
Pump
Conductivity Suppressor:
Chromatography Module
Detector:
Eluent Generator EG50:
Autosampler:
Columns :
Sample loop:
Load Volume:
Rinse Solution:
Eluent Flow:
Typical System Back-pressure:
Total analysis time:
Dionex DX500
GP40, 2-mm microbore
Dionex 2-mm Ultra II ASRS external water mode, 100 mA
Dionex LC30, temperature controlled at 35 °C
Dionex CD20 suppressed conductivity detector, background
conductivity: 1.0 [j,S
0.50, 65 and 100 mMNaOH (see Table IB)
Dionex AS40
Concentrator column Dionex Cryptand Cl, 4 x 35-mm
Guard column Dionex AG16, 2 x 50-mm
Analytical column Dionex AS 16, 2 x 250-mm
Cryptand Cl concentrator column (b)
2.0 mL of sample
l.OmLof IQmMNaOH
0.25 mL/min
2350 psi
43 minutes
(a) Mention of trade names or commercial products does not constitute endorsement or
recommendation for use.
(b) See Section 11.4
Standard Conditions and Equipment for Confirmation Analyses ^:
Ion Chromatograph:
Pump
Conductivity Suppressor:
Chromatography Module
Detector:
Eluent Generator EG50:
Autosampler:
Columns :
Sample loop:
Load Volume:
Rinse Solution:
Eluent Flow:
Typical System Back-pressure:
Total analysis time:
Dionex DX500
GP40, 2-mm microbore
Dionex 2-mm Ultra II ASRS external water mode, 100 mA
Dionex LC30, temperature controlled at 35 °C
Dionex CD20 suppressed conductivity detector, background
conductivity: 1.0 [j,S
0.50, 65 and 100 mMNaOH (see Table IB)
Dionex AS40
Concentrator column Dionex Cryptand Cl, 4 x 35-mm
Guard column Dionex AG20, 2 x 50-mm
Analytical column Dionex AS20, 2 x 250-mm
Cryptand Cl concentrator column (b)
2.0 mL of sample
l.OmLof IQmMNaOH
0.25 mL/min
2350 psi
48 minutes
(a) Mention of trade names or commercial products does not constitute endorsement or
recommendation for use.
(b) See Section 11.4
314.1-28
-------
EPA815-R-05-009
TABLE I.B. TIMING SEQUENCE FOR EPA METHOD 314.1 WITH AS16 and AS20
COLUMNS
Time
-13.60
-13.599
-5.010
-5.000
0.000
11.999
12.000
24.999
24.500
29.000
30.000
30.000
Eluent
100 mM
100 mM
100 mM
0.50 mM
0.50 mM
0.50 mM
0.50 mM
0.50 mM
65 mM
65 mM
100 mM
100 mM
100 mM
100 mM
Duration
0.01 sec
138 sec
1740 sec
Command
Pump Relay l.Open
Pump Relay 1. Closed
Start change to 0.50
mM
ChangetoO.SOmM
ECD.Autozero
ECD_l.Acqon
Pump Inject Valve. Inj
ectPosition
Start change to 65 mM
Change to 65 mM
Start change to 100
mM
Change to lOOmM
Pump Inject Valve. Lo
adPosition (1740 sec)
BCD l.Acqoff
Wait/end
Function
Initiates the program
Starts the AS40
ChangetoO.SOmM
Establish 0.50 mM in columns
Auto-zero the detector
Start data collection
Switch concentrator column to
inject position, elute perchl orate
off trap & refocus on AG16
Change to 65 mM
Separate & detect perchlorate
Change to lOOmM
Clean columns & establish
capacity of trap column
Switch concentrator column to
load position
Stop data collection
Eluent Generator ready to start
next run
Minor changes for AS20 columns
Increase the run time to 35 minutes to allow for the fact that perchlorate elutes about 4 minutes
later on the AS20 column.
314.1-29
-------
EPA815-R-05-009
TABLE 2. LOWEST CONCENTRATION MRL AND DLs FOR PERCHLORATE
Analytical
Column
AS16
AS20
Analyte
C1O4"
cicv
LCMRLa (ug/L)
0.14
0.13
*DL (ug/L)
0.03b
0.03 b
aLCMRLs were calculated according to the procedure in reference 1
*The DL was calculated from data acquired on a single day
bReplicate fortifications at 0.10 jig/L
TABLE 3. 1C PRECISION AND RECOVERY DATA FOR PERCHLORATE IN VARIOUS
MATRICES WITH AS16 COLUMNS (n=7)
Matrix
*Reagent Water
Chlorinated Surface Water
Chloraminated Surface Water
Chlorinated Ground Water
***LFSSM
Unfortified
Concentration
(ug/L)
<0.14**
<0.14**
0.63
0.63
<0.14**
<0.14**
<0.14**
<0.14**
<0.14**
<0.14**
Fortified
Concentration
(ug/L)
0.50
5.0
1.0
5.0
1.0
5.0
1.0
5.0
0.50
5.0
Mean %
Recovery
102
90.0
82.6
85.8
83.1
89.3
75.9
92.4
102
80.9
% RSD
2.6
3.2
2.7
2.0
3.6
1.8
5.4
3.3
2.8
1.3
* Reagent water containing 100 mg/L LSSM. **The LCMRL = 0.14 ug/L for the AS16 column.
***LFSSM Reagent water containing 1000 mg/L LSSM. Described in Section 3.11 and 3.13.
TABLE 4. 1C PRECISION AND RECOVERY DATA FOR PERCHLORATE IN VARIOUS
MATRICES WITH AS20 COLUMNS (n=7)
Matrix
*Reagent Water
Chloraminated Surface Water
Chlorinated Ground Water
**LFSSM
Unfortified
Concentration
(Ug/L)
<0.13**
<0.13**
<0.13**
<0.13**
0.22
0.22
<0.13**
<0.13**
Fortified
Concentration
(Ug/L)
0.50
5.0
0.50
5.0
0.50
5.0
0.50
5.0
Mean %
Recovery
104
94.2
108
97.8
96.2
98.0
97.4
86.3
% RSD
5.3
1.5
2.2
2.0
9.4
0.70
4.4
1.3
* Reagent water containing 100 mg/L LSSM. **The LCMRL = 0.13 ug/L for the AS20 column.
**LFSSM Reagent water containing 1000 mg/L LSSM. Described in Section 3.11 and3.13.
314.1-30
-------
EPA815-R-05-009
TABLE 5. INITIAL DEMONSTRATION OF CAPABILITY QUALITY CONTROL
REQUIREMENTS
Method
Reference
Requirement
Specification and
Frequency
Acceptance Criteria
Section
9.2.1
Demonstration of
Low System
Background
Analyze a LRB and
LFSSMB prior to any other
IDC steps.
Demonstrate that perchlorate is
below 1/3 of the MRL and that
possible interferences from
sampling protocols do not
prevent the identification and
quantification of perchlorate.
Section
9.2.4
Minimum
Reporting Limit
(MRL)
Confirmation
Fortify and analyze 7
replicate LFBs at the
proposed MRL
concentration. Calculate
the mean and the Half
Range (FIR). Confirm that
the Upper PIR and Lower
PIR (Sect. 9.2.4) meet the
recovery criteria.
Section 9.2.4.2
Upper PIR < 150%.
Lower PIR > 50%.
Section
9.2.2
Demonstration of
Precision
Analyze 7 replicate LFBs
fortified near the mid-point
of the calibration curve
%RSDmustbe < 20%.
Section
9.2.3
Demonstration of
Accuracy)
Calculate average recovery
for replicates used in
Section 9.2.3.
Mean recovery + 25% of true
value.
Section
9.2.5
Validation of MRL
in 1000 mg/L
LFSSM
Analyze 7 replicate
LFSSMs fortified at the
MRL.
Section 9.2.4.2
Upper PIR < 150%.
Lower PIR > 50%.
Section
9.4.1
Quality Control
Sample
During IDC, each time a
new analyte PDS is made,
every time the instrument
is calibrated and at least
quarterly.
The result for perchlorate must
be 75-125% of the true value.
314.1-31
-------
EPA815-R-05-009
TABLE 6. ONGOING QUALITY CONTROL REQUIREMENTS (SUMMARY)
Method
Reference
Requirement
Specification and
Frequency
Acceptance Criteria
Section
10.2
Initial Calibration
Use external standard
calibration technique to
generate a first or second
order calibration curve. Use
at least 5 standard
concentrations.
Check the calibration curve
as described in Section 10.2.
Analyze a QCS near the mid-
point of the calibration
curve.
When each calibration standard
is calculated as an unknown
using the calibration curve, the
result should be:
Level
Result
< MRL + 50% True value
> MRL + 25% True value
to high CAL
The result for perchlorate must
be 75-125% of the true value.
Recalibration is recommended
if these criteria are not met.
Section
9.3.2
Laboratory
Synthetic Sample
Matrix Blank
(LSSMB)
Daily, or with each Analysis
Batch of up to 20 field
samples, whichever is more
frequent.
Demonstrate that the
perchlorate is below V3 the
MRL, and confirm that possible
interferences do not prevent
quantification of perchlorate. If
the target exceeds V3 the MRL,
the results for perchlorate in the
Analysis Batch are invalid.
Section
9.3.3
Continuing
Calibration Check
(CCC) Standards
Verify initial calibration by
analyzing a low-level CCC at
the beginning of each
Analysis Batch. Subsequent
CCCs are required after
every 10 field samples, and
after the last field sample in a
batch.
Low CCC - at or below the
MRL concentration
Mid CCC - near midpoint in
calibration curve
High CCC - near the highest
calibration standard.
For each CCC the result must
be
CCC Level
Result
< MRL + 50% True value
> MRL + 25% True value
to high CAL
314.1-32
-------
EPA815-R-05-009
TABLE 6. (Continued)
Method
Reference
Requirement
Specification and
Frequency
Acceptance Criteria
Section
9.3.4
Laboratory
Fortified Synthetic
Sample Matrix
CCCs
(LFSSM CCC)
In order to monitor trapping
efficiency during an Analysis
Batch, the CCC standards,
prepared in the LFSSM
(Sect. 9.3.3) are also required
at the same frequency and
concentrations.
For each LFSSM CCC the
result must be
CCC Level Result
< MRL + 50% True value
> MRL + 25% True value
to high CAL
Section
9.3.6
Laboratory
Fortified Sample
Matrix (LFSM)
Analyze one LFSM per
Analysis Batch (20 field
samples or less). Fortify the
LFSM with perchlorate at a
concentration close to but
greater than the native
concentration (if known).
Calculate LFSM recoveries.
Recoveries for the LFSM must
be calculated (Sect. 9.3.6.3 ).
The result must be
LFSM Level Result
< MRL + 50% True value
> MRL + 25% True value
to high CAL
Section
9.3.7
Laboratory
Duplicate (LD) or
Laboratory
Fortified Sample
Matrix Duplicate
(LFSMD)
Analyze at least one LD or
LFSMD daily, or with each
Analysis Batch (20 samples
or less), whichever is more
frequent.
Precision must be calculated
(Sect. 9.3.7.2). The result must
be
Level Result
<2xMRL < 50%RPD
2xMRL
to high CAL
< 25%RPD
Section
9.4.1
Quality Control
Sample (QCS)
During IDC, each time a new
analyte PDS is made, every
time the instrument is
calibrated and at least
quarterly.
Results must be + 25% of the
expected value.
Section
Sample Holding
Time
28 days when processed and
stored according to sections
8.1 and 8.2 with appropriate
preservation and storage.
Sample results are valid only if
samples are extracted within
sample holding time.
314.1-33
-------
EPA815-R-05-009
TABLE 7. SAMPLE ANALYSIS BATCH WITH QC REQUIREMENTS
Injection
#
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
Sample
Description
Laboratory Synthetic Sample Matrix Blank (LSSMB)
Low-CCC at the MRL (0.5 ng/L)
Low-Laboratory Fortified Synthetic Sample Matrix CCC
(LFSSM CCC @ 0.5 ng/L)
Sample 1
Sample 2
Sample 2 - Laboratory Fortified Sample Matrix (LFSM)
Sample 2 - Laboratory Fortified Sample Matrix Duplicate
(LFSMD)
Sample 3
Sample 4
Sample 5
Sample 6
Sample 7
Sample 8
Sample 9
Sample 10
Mid-CCCat5.0ng/L
Mid-Laboratory Fortified Synthetic Sample Matrix CCC
(LFSSMCCC@5.0ng/L)
Sample 1 1
Sample 12
Acceptance
Criteria
< 1/3 MRL
0.25 to 0.75 ng/L
0.25 to 0.75 ng/L
sample analysis
sample analysis
Recovery of 75 - 125%
%RPD = + 25%
sample analysis
sample analysis
sample analysis
sample analysis
sample analysis
sample analysis
sample analysis
sample analysis
3.75-6.5 |ig/L
3.75-6.5 |ig/L
sample analysis
sample analysis
CONTINUED on NEXT PAGE
314.1-34
-------
TABLE 7. (Continued)
EPA815-R-05-009
Injection
#
20
21
22
23
24
25
26
27
28
29
Sample
Description
Sample 13
Sample 14
Sample 15
Sample 16
Sample 17
Sample 18
Sample 19
Sample 20
High-CCC at 10 ng/L
High-Laboratory Fortified Synthetic Sample Matrix CCC
(LFSSMCCC@ 10ng/L)
Acceptance
Criteria
sample analysis
sample analysis
sample analysis
sample analysis
sample analysis
sample analysis
sample analysis
sample analysis
7.5 - 12.5 ng/L
7.5 - 12.5 ng/L
314.1-35
-------
EPA815-R-05-009
Figure 1
EPA METHOD 314.1 CHROMATOGRAM of 5.0 fj.g/L C1O4 and 300 ^g/L 4-C1BSA WITH
IONPAC AS16 AND AS20 COLUMNS
AS16 CIO4- Recovery = 79.0%
AS20 CIO4- Recovery = 94.0%
-0.10
0.0 2.0 4.0 6.0 8.0 10.0 12.0 14.0 16.0 18.0 20.0 22.0 24.0 26.0 28.0 30.0 32.0 35.0
314.1-36
-------
EPA815-R-05-009
s.a
2.5
2.0
1.5
1.0
-O.I
-O.i
Figure 2
1C CHROMATOGRAM of SURFACE and GROUND WATER
FORTIFIED WITH 1.0 u.g/L C1O4
Chlorinated surface water 1.6 ug/L
Native level = 0.60 ug/L CIO4-
Chlorinated ground water 1.1 ug/L
Native level =
-------
EPA815-R-05-009
Figure 3
1C CHROMATOGRAM of 3.0 ng/L C1O4 in 50, 500 and 1000 mg/L LSSM WITH the IONPAC
AS16 COLUMN
4.0-
3.0
2.0'
MS
3.0 |jg/L CICV in 1000 mg/L LSSM
peak area = 0.0611 uS*min
3.0 |jg/L CIO4' in 500 mg/L LSSM
peak area = 0.0660 uS*min
3.0 |jg/L CIO4" in 50 mg/L LSSM
peak area = 0.0666 uS*min
-0.201 > ' ' | > ' ' | ' < ' | ' < ' | ' ' ' | ' ' > | ' ' < | ' ' ' | 'T ' i ' ' ' | ' ' ' | < ' ' | < ' ' | ' ' ' | '
0.0 2.0 4.0 6.0 8.0 10.0 12.0 14.0 16.0 18.0 20.0 22.0 24.0 26.0 28.0 30.0
314.1-38
------- |