&EPA
METHOD 302.0: DETERMINATION OF BROMATE IN
DRINKING WATER USING TWO-DIMENSIONAL ION
CHROMATOGRAPHY WITH SUPPRESSED
CONDUCTIVITY DETECTION
Office of Water (MLK 140) EPA Document No. 815-B-09-014  September 2009  www.epa.gov/safewater

-------
METHOD 302.0   DETERMINATION OF BROMATE IN DRINKING WATER USING
                 TWO-DIMENSIONAL ION CHROMATOGRAPHY WITH SUPPRESSED
                 CONDUCTIVITY DETECTION
                                    Version 1.0
                                  September 2009
Herbert P. Wagner (Lakeshore Engineering Services, Inc.)
Barry V. Pepich (Shaw Environmental, Inc.)
Chris Pohl, Kan nan Srinivasan, Brian De Borba, Rong Lin (Dionex Corp.)
David J. Munch (U.S. EPA, Office of Ground Water and Drinking Water)
                          TECHNICAL SUPPORT CENTER
               OFFICE OF GROUND WATER AND DRINKING WATER
                  U. S. ENVIRONMENTAL PROTECTION AGENCY
                             CINCINNATI, OHIO 45268
                                      302.0-1

-------
                                      METHOD 302.0

 DETERMINATION OF BROMATE IN DRINKING WATERS USING TWO-DIMENSIONAL
      ION CHROMATOGRAPHY WITH SUPPRESSED CONDUCTIVITY DETECTION

1.  SCOPE AND APPLICATION

   1.1   This is a large volume (1.0 mL), two-dimensional (2-D) ion chromatographic (1C) method
         using suppressed conductivity detection for the determination of bromate in raw and finished
         drinking waters.  Because this method utilizes two dissimilar 1C columns it does not require
         second column confirmation. Detection and  quantitation in the second dimension are
         accomplished by suppressed conductivity detection.  Precision and accuracy data have been
         generated for bromate using this 2-D 1C method in reagent water, finished groundwater,
         finished surface water and a Laboratory Fortified Synthetic Sample Matrix (LFSSM). The
         single laboratory Lowest Concentration Minimum Reporting Level (LCMRL) has also been
         determined in reagent water.l

                                                     Chemical Abstract Services
            Analvte                                  Registry Number (CASRN)
            Bromate                                           15541-45-4
   1.2   The Minimum Reporting Level (MRL) is the lowest analyte concentration that meets Data
         Quality Objectives (DQOs) that are developed based on the intended use of this method.  The
         single laboratory LCMRL is the lowest true concentration for which the future recovery is
         predicted to fall between 50 and 150 percent recovery with 99% confidence.  The single
         laboratory LCMRL for bromate was 0.18 ug/L using a 1.0-mL injection volume (see Table 2).
         The procedure used to determine the LCMRL is described elsewhere.l

   1.3   Laboratories using this method will not be required to determine the LCMRL, but will need to
         demonstrate that their laboratory MRL for this method meets the requirements described in
         Section 9.2.4.

   1.4   Detection Limit (DL) is defined as the statistically calculated minimum concentration that can
         be measured with 99% confidence that the reported value is greater than zero.  The DL for
         bromate is dependent on sample matrix, fortification concentration, and instrument
         performance.  Determining the DL for bromate in this method is optional (Sect. 9.2.6).  The
         reagent water  DL for bromate using a 1.0-mL injection volume was calculated to be 0.12 |ig/L
         using 7 reagent water (RW) replicates fortified at 0.20 |ig/L (see Table 2).

   1.5   This method is intended for use by analysts skilled in the operation of 1C instrumentation, and
         the interpretation of the associated data.

   1.6   METHOD FLEXIBILITY - In recognition of technological advances in analytical systems
         and techniques, the laboratory is permitted to modify the separation technique, 1C columns,
         concentrator column and mobile phase composition. However, any modifications must
         maintain the basic chromatographic elements of this new technique.  This includes the initial

                                           302.0-2

-------
         separation of analytes on a primary 1C column that are heart-cut after suppression and directed
         to a concentrator column. After concentration, analytes must then be eluted onto a second,
         dissimilar 1C column for separation and quantitation. For high sensitivity applications, the
         system should have a second-dimension column with a lower cross sectional area, which
         operates at a relatively lower flow rate in order to achieve sensitivity enhancement
         proportional to the ratio  of the flow rate reduction. Method modifications should be
         considered only to improve method performance. Modifications that are introduced in the
         interest of reducing cost or sample processing time, but result in poorer method performance
         (see Sect. 9.4 for criteria), may not be used.  In all cases where method modifications are
         proposed, the analyst must first redetermine the cut window and establish an acceptable
         calibration.  Then the analyst must demonstrate acceptable method performance by
         conducting the procedures outlined in the initial demonstration of capability (IDC, Sect. 9.2),
         verifying that all ongoing QC acceptance criteria can be routinely met (Tables 4 and 5),  and
         properly documenting the method changes (Sect. 9.4). Changes may not be made to sample
         collection and preservation (Sect. 8) or to the quality control requirements (Sect. 9).

2. SUMMARY OF METHOD

   2.1   Water samples are collected in the field using normal  collection techniques and dechlorinated
         with ethylenediamine (EDA). A 1.0-mL sample  aliquot is injected onto a 4-mm 1C  column.
         Separation of bromate is achieved in the first dimension (1-D) using 10 mM KOH at a flow
         rate of 1.0 mL per minute. Approximately 2 mL of the suppressed eluent containing the
         bromate is diverted from the first dimension column to a concentrator column used in place of
         the sample loop of the second dimension (2-D) injection valve. The concentrator column has
         low backpressure but sufficient capacity to trap the bromate ions quantitatively in the
         suppressed eluent. In this manner, bromate is separated from other matrix ions and
         concentrated on a trapping column. The heart-cut portion of the 1-D chromatogram is eluted
         off the concentrator column and onto a smaller diameter (2 mm diameter) guard and analytical
         column that have different selectivity from the first dimension columns to facilitate the 2-D
         separation using 10 mM KOH at a flow rate of 0.25 mL per minute.  Bromate is quantitated
         using the external standard method.

         There are several advantages to this version of the method. This version is compatible with a
         large  sample injection volume (up tol.O-mL). This is due to the high capacity of the 1-D
         analytical column and its higher selectivity for bromate relative to the other matrix ions. The
         suppressed eluent from the first dimension column, which is essentially water containing the
         anions of interest and a subset of the matrix interferences that co-elute in the heart-cut
         window, are efficiently retained and refocused on the concentrator column.  Because the
         second dimension is operated at a lower flow rate relative to the first dimension and uses a
         smaller cross sectional area column, this version  of the method offers enhanced sensitivity.
         The increase in sensitivity of the present setup is directly proportional to the flow rate ratio of
         the first dimension to the second dimension columns.   This yields method sensitivity
         comparable to the EPA bromate methods that utilize the addition of postcolumn reagents
         (317.0 and 326.0) and UV/Vis  detection. Finally, the 2-D 1C method combines two columns
         with different selectivity thereby eliminating the  need for second column confirmation.
                                            302.0-3

-------
3.  DEFINITIONS

   3.1    ANALYSIS BATCH - A sequence of field samples, which are analyzed within a 24-hour
          period and include no more than 20 field samples. An Analysis Batch must also include all
          required QC samples, which do not contribute to the maximum field sample total of 20.  For
          this method, the required QC samples include:

               Laboratory Reagent Blank (LRB)
               Continuing Calibration Check (CCC) Standards
               Laboratory Fortified Sample Matrix (LFSM)
               Laboratory Duplicate (LD) or a Laboratory Fortified Sample Matrix Duplicate
               (LFSMD)
               Laboratory Fortified Synthetic Sample Matrix CCC Standard (LFSSM CCC)

   3.2    ANALYTE FORTIFICATION SOLUTIONS (AFS) - The Analyte Fortification Solutions
          are prepared by dilution of the Analyte Secondary Dilution Solutions (SDS) and are used to
          fortify the LFSMs and the LFSMDs with bromate. It is recommended that multiple
          concentrations be prepared so that the fortification levels can be rotated or adjusted to the
          concentration of target analyte in the native samples.

   3.3    CALIBRATION STANDARD (CAL) - A solution of the target analyte prepared from the
          Bromate Primary Dilution Solution or Bromate Stock Standard Solution. The CAL solutions
          are used to calibrate the instrument response with respect to analyte concentration.

   3.4    CONTINUING CALIBRATION CHECK STANDARD (CCC) - A calibration check
          standard containing the method analyte which is analyzed periodically throughout an
          Analysis Batch, to verify the accuracy of the existing calibration for that analyte.

   3.5    DETECTION LIMIT (DL) - The minimum concentration of an analyte that can be identified
          measured and reported with 99% confidence that the analyte concentration is greater than
          zero. This is a statistical determination (Sect. 9.2.6), and accurate quantitation is not
          expected at this level.2

   3.6    LABORATORY DUPLICATE (LD) - Two sample aliquots (LDi and LD2), from a single
          field sample bottle, and analyzed separately with identical procedures. Analyses of LDi and
          LD2 indicate precision associated specifically with laboratory procedures by removing
          variation contributed from sample collection and storage procedures.

   3.7    LABORATORY FORTIFIED BLANK (LFB) - An aliquot of reagent water or other blank
          matrix to which a known quantity of the method analyte is added. The LFB is analyzed
          exactly like a sample.  Its purpose is to determine whether the methodology is in control, and
          whether the laboratory is capable of making accurate and precise measurements.

   3.8    LABORATORY FORTIFIED SAMPLE MATRIX (LFSM) - An aliquot of a field sample to
          which a known quantity of the method analyte is added. The LFSM is processed and
          analyzed exactly like a field sample, and its purpose is to determine whether the field sample
          matrix contributes bias to the analytical results.  The background concentration of the analyte
                                           302.0-4

-------
      in the field sample matrix must be determined in a separate aliquot and the measured value in
      the LFSM corrected for the native concentration.

3.9    LABORATORY FORTIFIED SAMPLE MATRIX DUPLICATE (LFSMD) - A second
      aliquot of the field sample used to prepare the LFSM, which is fortified and analyzed
      identically to the LFSM.  The LFSMD is used instead of the Laboratory Duplicate to assess
      method precision and accuracy when the occurrence of the target analyte is infrequent.

3.10  LABORATORY FORTIFIED SYNTHETIC SAMPLE MATRIX (LFSSM) - An aliquot of
      the LSSM (Sect. 7.2.2) which is fortified with bromate. The LFSSM is used to set the start
      time for the cut window in the 1-D during the initial demonstration of capability (IDC) (Sect.
      9.2). The LFSSM is also used to determine the precision and accuracy of the method during
      the IDC (Sects. 9.2.2 and 9.2.3). The LFSSM samples are treated like the CCCs.

3.11  LABORATORY FORTIFIED SYNTHETIC SAMPLE MATRIX CONTINUING
      CALIBRATION CHECK STANDARD (LFSSM CCC) - An aliquot of the LSSM (Sect.
      7.2.2) which is fortified with bromate at a concentration equal to one of the CCCs. In this
      method, a LFSSM CCC at a concentration equal to the highest calibration level should be
      analyzed near the beginning of each Analysis Batch (Sect. 9.3.3) to confirm that the 1-D
      heart cutting procedure has acceptable recovery in high inorganic matrices.

3.12  LABORATORY REAGENT BLANK (LRB) - An aliquot of reagent water or other blank
      matrix that is treated exactly as a sample including exposure to storage containers. The LRB
      is used to determine if the method analyte or other interferences are present in the laboratory
      environment, the reagents, or the apparatus.

3.13  LABORATORY SYNTHETIC SAMPLE MATRIX (LS SM) - An aliquot of reagent water
      that is fortified with 100 mg/L of the sodium salts of chloride, bicarbonate and sulfate and 10
      mg/L phosphate as phosphorus and nitrate as nitrogen using the sodium salts of phosphate
      and nitrate.

3.14  LOWEST CONCENTRATION MINIMUM REPORTING LEVEL (LCMRL) - The single-
      laboratory LCMRL is the lowest true concentration for which the future recovery is predicted
      to fall between 50  and 150 percent recovery with 99% confidence. l

3.15  MATERIAL SAFETY DATA SHEET (MSDS) - Written information provided by vendors
      concerning a chemical's toxicity,  health hazards, physical properties, fire, and reactivity data
      including storage,  spill, and handling precautions.

3.16  MINIMUM REPORTING LEVEL (MRL) - The minimum concentration that can be
      reported by a laboratory as a quantified value for the target analyte in a sample following
      analysis. This defined concentration must meet the criteria defined in  Section 9.2.4 and must
      be no lower than the concentration of the lowest calibration standard for the target analyte.

3.17  PRIMARY DILUTION STANDARD SOLUTION (PDS) - A solution containing the
      method analyte prepared in the laboratory from stock standard solutions and diluted as
      needed to prepare calibration solutions and other analyte-containing solutions.

                                       302.0-5

-------
   3.18   QUALITY CONTROL SAMPLE (QCS) - A solution containing the method analyte at a
          known concentration that is obtained from a source external to the laboratory and different
          from the source of calibration standards.  The QCS is used to verify the accuracy of the
          calibration standards and the integrity of the calibration curve.

   3.19   REAGENT WATER (RW) - Purified water which does not contain any measurable quantity
          of the target analyte or interfering compounds at or above 1/3 the MRL.

   3.20   SECONDARY DILUTION STANDARD SOLUTION (SDS) - A solution containing the
          method analyte prepared in the laboratory from the PDS and diluted as needed to prepare
          calibration solutions and other analyte solutions.

   3.21   STOCK STANDARD SOLUTION (SSS) - A concentrated solution containing the method
          analyte prepared in the laboratory using assayed reference materials or purchased from a
          reputable commercial source, so that the concentration and purity of analytes are traceable to
          certificates of analysis.

4.  INTERFERENCES

   4.1   Interferences can be divided into three different categories:  (i) direct chromatographic co-
         elution, where an analyte response is observed at very nearly the same retention time (RT) as
         the target analyte; (ii) concentration dependant co-elution, which is observed when the
         response of higher than typical concentrations of the neighboring peak overlaps into the
         retention window of the target analyte; and (iii) ionic character displacement, where retention
         times may significantly shift due to the influence of high ionic strength matrices (high mineral
         content or total dissolved solids) overloading the exchange sites on the column and
         significantly shortening the target  analyte's retention time.

       4.1.1  A direct chromatographic co-elution may be solved by changing column selectivity in
             one or both dimensions of the 2-D 1C method, adjusting eluent strength in one or both
             dimensions, modifying the eluent with organic solvents (if compatible with 1C columns),
             changing the detection system, or selective removal of the interference with sample
             pretreatment.  Sample dilution will have little to no effect on direct chromatographic co-
             elution.  The analyst must verify that any change made to the chromatographic
             parameters does not induce any negative affects on method performance by repeating and
             passing all the QC criteria as  described  in Section 9.2.

       4.1.2  Sample dilution may resolve some of the difficulties if the interference is the result of
             either concentration dependant co-elution or ionic character displacement, but it must be
             clarified that sample dilution will alter the MRL by a proportional factor equivalent to
             that of the dilution. Therefore, careful consideration of Data Quality Objectives (DQOs)
             should be given prior to performing such a dilution.

   4.2   Method interferences may be caused by contaminants in solvents, reagents (including reagent
         water), sample bottles and caps, and other sample processing hardware that lead to discrete
         artifacts and/or elevated baselines in the chromatograms. All items such as these must be
         routinely demonstrated to be free from interferences (less than Vs the bromate MRL) by

                                             302.0-6

-------
         analyses of an LRB as described in Section 9.2.1. Subtracting blank values from sample
         results is not permitted.

   4.3   Matrix interferences may be caused by contaminants that are present in the sample. The
         extent of matrix interferences will vary considerably from source to source, depending upon
         the nature of the water.  Water samples high in organic carbon or total dissolved solids may
         have elevated baselines  or interfering peaks.

       4.3.1  Equipment and containers used for sample collection and storage have the potential to
             introduce interferences. The potential for interferences from these sources must be
             investigated during the IDC (Sect. 9.2) by preparing and analyzing a LRB. This
             procedure should be repeated each time that a new brand and/or lot of materials are used
             to ensure that contamination does not hinder analyte identification and quantitation.

   4.4   When first-dimension heart-cut windows are properly set, this method demonstrates adequate
         performance in water matrices that contain up to 100 mg/L of each common anion (sulfate,
         bicarbonate, and chloride), and phosphate as phosphorus and nitrate as nitrogen at 10 mg/L.
         While this addresses the majority of source and finished drinking waters, there is a possibility
         that water matrices could exceed this level.  The  analyst should monitor all first dimension
         chromatograms (Sect. 11.3.4) to confirm that sample matrix does not overload the primary
         column capacity and require dilution.

5.  SAFETY

   5.1   The toxicity or carcinogenicity of each reagent used in this method has not been precisely
         defined. Each chemical should be treated as a potential health hazard, and exposure to these
         chemicals should be minimized. Each laboratory is responsible for maintaining an awareness
         of OSHA regulations regarding safe handling of chemicals used in this method. A reference
         file of MSDSs should be made available to all personnel  involved in the chemical analysis.
         Additional references to laboratory safety are available.3"5

6.  EQUIPMENT AND SUPPLIES  (References to specific brands or catalog numbers are included
   for illustration only and do not imply endorsement of the product.  This does not preclude the use of
   other vendors, supplies or configurations.)

   6.1   SAMPLE CONTAINERS - 125-mL brown  Nalgene bottles (Fisher Cat. No. 03-313-3C or
         equivalent).

   6.2   VOLUMETRIC FLASKS - Class A, suggested sizes include 10, 50, 100, 250, 500 and 1000
         mL for preparation of standards and eluents.

   6.3   GRADUATED CYLINDERS - Suggested sizes include  25 and 1000 mL.

   6.4   AUTO PIPETTES - Capable of delivering variable volumes from 1.0 to 2500 uL.

   6.5   ANALYTICAL BALANCE - Capable of weighing to the nearest 0.0001 g.
                                            302.0-7

-------
6.6   DUAL ION CHROMATOGRAPHY SYSTEM WITH SUPPRESSED CONDUCTIVITY
      DETECTION (1C) - This section describes the instrument configuration that was used to
      collect the data in Section 17.  A Dionex Model ICS-3000 Dual system, consisting of a Dual
      Pump (DP) module, Eluent Generator (EG) module, Detector/Chromatography (DC) module,
      and Autosampler (AS), was used to collect the data presented in this method. Equivalent
      systems may be used The 1C system should also have a temperature controlled column
      compartment and be capable of operating above room temperature (30 °C) and include dual 1C
      pumps and all required accessories, including guard, analytical, and concentrator columns,
      detector/chromatography module, dual eluent generators, continuously-regenerated anion trap
      columns, compressed gases, autosampler, suppressors, carbonate removal devices (CRD),
      dual conductivity detectors, and a computer-based data acquisition and control system.
      Additionally, the system must be capable of performing automated, two-dimensional 1C,
      including performing inline column concentration and matrix elimination steps as described in
      Section 2.1. A schematic diagram  of the instrumentation for this 2-D 1C method is shown in
      Figure 1.  Table  1 provides full details of the instrumental conditions.

   6.6.1   DUAL PUMP MODULE - A DP Dual Gradient-Gradient Pumping Module with dual
          channel degas devices (Dionex DP, P/N 061712), was used to generate the data for this
          method. Equivalent modules may be used. The dual pump system used for method
          development was capable of supplying a flow rate of approximately 1.0 mL/min to the
          first dimension column and approximately 0.25 mL/min to the second dimension column.

   6.6.2   ELUENT GENERATOR MODULE - A dual channel EG Module (Dionex EG, P/N
          061714) with dual potassium hydroxide cartridges (EluGen® Cartridge, EGC, P/N
          058900) was used to prepare the potassium hydroxide eluent for both the first and second
          dimensions  of this method. An equivalent eluent generator may be used and/or manually
          prepared eluents may also be used provided that adequate resolution, peak shape,
          capacity, accuracy, and precision (Sect. 9.2) are obtained. Care must be exercised with
          manually prepared hydroxide eluents to prevent formation of carbonate in the eluent from
          exposure to the atmosphere, which can dramatically alter the chromatography and affect
          sensitivity.

       6.6.2.1 CONTINUOUSLY REGENERATED ANION TRAP COLUMNS - 1C eluent
             purification columns (Dionex CR-ATC, P/N 060477 or equivalent). Any in-line,
             resin-based manual or electrolytic trapping column that provides adequate eluent
             purification for ultra trace  analysis and performance (Sect. 9.2) may be used. A CR-
             ATC device, or equivalent, was used for eluent purification for both the first and
             second dimension eluents.

             NOTE:  For the configuration in Figure 1, the pump and eluent generator modules in
             combination must be capable of delivering different isocratic eluent concentrations to
             the columns in the first and second dimension. The same requirement applies for
             manually prepared eluents with a dual pumping system; the pump system must be
             capable  of delivering two different, independent isocratic concentrations of eluents to
             the first  and second dimension  columns. In addition, the system should also be
             capable  of providing a step isocratic eluent concentration change or a controlled
             gradient change to both dimensions independently.  This allows the first and second
             dimension columns to be cycled to a higher eluent concentration in order to clean
                                        302.0-8

-------
          residual matrix components from the columns prior to introduction of the next
          sample. This is important to ensure maximum column life and to minimize potential
          carryover and/or interferences from one sample to the next.

6.6.3  DETECTOR/CHROMATOGRAPHY MODULE - A DC Module (Dionex DC, P/N
      061793 or equivalent) equipped with dual injection valves and capable of maintaining the
      columns and the suppressors at 30 °C, and conductivity cell at 35 °C is recommended.

      NOTE: For optimal performance of this system, the conductivity cell should be set at a
      higher temperature than the analytical columns. For example, if the columns are set at 30
      °C, the cell should be set at 35  °C.

6.6.4  FIRST DIMENSION GUARD COLUMN - An 1C column, 4 x 50-mm (Dionex
      IonPac®AG19, P/N 062887 or equivalent).  Any guard column that provides adequate
      protection for the analytical column and does not have an adverse effect on the peak
      shape may be used.

6.6.5  FIRST DIMENSION ANALYTICAL COLUMN - An 1C column, 4 x 250 mm (Dionex
      IonPac®AS19, P/N 062885 or equivalent). Any analytical column that provides adequate
      resolution, peak shape, capacity, accuracy, and precision  (Sect. 9.2) may be used. The
      separation mechanism for the first dimension analytical column must differ from the
      second dimension column in selectivity.

6.6.6  FIRST DIMENSIOIN ANION SUPPRESSOR DEVICE  - An 1C eluent suppression
      device, 4 mm (Dionex Anion Self-Regenerating Suppressor, ASRS Ultra II, P/N 061561
      or equivalent). An equivalent in-line suppression device  that effectively suppresses the
      conductance of the eluent and converts the analyte to the  hydronium form prior to
      conductivity detection, as well as provides adequate efficiency, resolution, peak shape,
      capacity, accuracy, precision, and a comparable MRL and DL (Section 9.2) may be used.
      Adequate baseline stability should be attained as  measured by a baseline noise of no more
      than 5  nS per minute over the background conductivity.  The first dimension suppressor
      must be compatible with the first dimension guard and analytical column.

      NOTE: The  conductivity suppressor was set to perform  electrolytic suppression  at a
      current setting of 161 mA. It was important to operate the suppressor in the external
      water mode to reduce baseline noise and achieve optimal  method performance.

6.6.7  FIRST DIMENSION CARBONATE REMOVAL DEVICE - An 1C carbonate removal
      device, 4 mm (Dionex CRD, P/N 062983 or equivalent).  Any in-line carbonate removal
      device that effectively removes the carbonate peak from the suppressed eluent stream
      prior to conductivity detection  and provides adequate efficiency,  resolution, peak shape,
      capacity, accuracy, and precision for bromate (Section 9.2) may be used.  The first
      dimension CRD must be compatible with (e.g., 4 mm in this example) the first dimension
      guard and analytical column.

6.6.8  FIRST DIMENSION CONDUCTIVITY DETECTOR -  A Conductivity detector and
      integrated cell (Dionex CD P/N 061716, or equivalent) capable of providing data as
      required in Section 9.2.  A Standard Bore Temperature Stabilizer (0.010 inch ID, Dionex
                                    302.0-9

-------
      P/N 062561), was also used to equilibrate the temperature of the eluent to that of the first
      dimension guard and analytical column. Equivalent stabilizers may be used.

      NOTE:  The conductivity detector cell temperature should be controlled at a temperature
      above that of the analytical column. For method development, the conductivity detector
      was set at 35 °C to minimize bubble formation and condensation between analytical
      column, suppressor and CRT) and to stabilize the temperature of the detector cell itself.

6.6.9  CONCENTRATOR COLUMN - An 1C trapping column, 5 x 23 mm (Dionex UTAC-
      ULP1, P/N 063475 or equivalent). Any concentrator column that provides effective
      retention/trapping and release of bromate while providing the resolution, peak shape,
      capacity, accuracy, and precision (Sect. 9.2) may be used. The concentrator column
      should not release sulfonated leachates that would affect the quantitation of bromate.

   6.6.9.1 Alternate concentrator columns are allowed, but prior to their use, they must be
          evaluated to determine the first dimension cut window (Sect. 10.2.2). They must be
          determined to have sufficient capacity to quantitatively trap bromate in the LFSSM
          CCC (Sect. 10.3.3) and should have relatively low backpressure since the
          concentrator column is placed as a post-suppressor device.

6.6.10 SECOND DIMENSION GUARD COLUMN - An 1C column, 2 x 50 mm (Dionex
      IonPac®AG24, P/N 064151 or equivalent). Any guard column that provides adequate
      protection for the analytical column and does not have an adverse effect on the peak
      shape may be used.

6.6.11 SECOND DIMENSION ANALYTICAL COLUMN - An 1C column, 2 x 250 mm
      (Dionex lonPac® AS24, P/N 064153 or equivalent). Any analytical column that provides
      adequate resolution, peak shape, capacity, accuracy, and precision (Sect. 9.2) may be
      used. The separation mechanism for the second dimension analytical column must differ
      from the first dimension column.

6.6.12 SECOND DIMENSIOIN ANION SUPPRESSOR DEVICE - An 1C eluent suppression
      device, 2 mm (Dionex Anion Self-Regenerating Suppressor, ASRS Ultra II, P/N 061562
      or equivalent). An equivalent in-line suppression device that effectively suppresses the
      conductance of the eluent prior to conductivity detection, and that provides adequate
      efficiency, resolution, peak shape, capacity,  accuracy, precision,  and a comparable MRL
      and DL (Section 9.2) may be used. Adequate baseline stability should be attained as
      measured by a baseline noise of no more than 5 nS per minute over the background
      conductivity. The second-dimension suppressor must be compatible with (e.g., 2 mm in
      this example) the second-dimension guard and analytical column.

      NOTE:  The conductivity suppressor was set to perform electrolytic suppression at a
      current setting of 41 mA. It was important to operate the suppressor in the external water
      mode to reduce baseline noise and achieve optimal method performance.

6.6.13 SECOND DIMENSION CARBONATE REMOVAL DEVICE - An 1C carbonate
      removal device (Dionex CRD, P/N 062986 or equivalent).  Any in-line carbonate
      removal device that effectively removes the carbonate peak from the suppressed eluent
                                     302.0-10

-------
             stream prior to conductivity detection of the method analyte and provides adequate
             efficiency,  resolution, peak shape, capacity, accuracy, and precision (Section 9.2) may be
             used. The second dimension CRD must be compatible with the second dimension guard
             and analytical column.

       6.6.14 SECOND DIMENSION CONDUCTIVITY DETECTOR - A Conductivity detector and
             integrated cell (Dionex CD P/N 061716, or equivalent) capable of providing data as
             required in  Section 9.2. A Microbore Temperature Stabilizer (0.005 inch ID, Dionex P/N
             062562, or  equivalent), was also used to equilibrate the temperature of the eluent to that
             of the second dimension guard and analytical column.

             NOTE: The conductivity detector cell temperature should be controlled at a temperature
             above the analytical column. For method development, the conductivity was set at 35 °C
             to minimize bubble formation and condensation between analytical column, suppressor
             and CRD and to stabilize the temperature of the detector cell itself.

       6.6.15 AUTOSAMPLER MODULE - An AS Autosampler Module with simultaneous injection,
             a sample prep option, and a large volume sample needle assembly (Dionex AS P/N
             063105) was used to generate data for this method. Any autosampler capable of
             automatically injecting up to 1.0 mL of sample may be used.

       6.6.16 DATA SYSTEM - An interfaced data system such as Dionex, Chromeleon Version 6.7
             (or equivalent) is required to acquire, store, and output conductivity data. The computer
             software should have the capability of processing stored conductivity data by recognizing
             and integrating a peak within a given retention time window. The software  should be
             capable of constructing linear regressions or quadratic calibration curves, and calculating
             analyte concentrations using the calibrations.

7.    REAGENTS AND STANDARDS

    7.1   REAGENTS - Reagent grade or better chemicals should be used in all tests. Unless
          otherwise indicated, it is intended that all reagents will conform to the specifications of the
          Committee on Analytical Reagents of the American Chemical Society (ACS), where such
          specifications are available. Other grades may be used, provided it is first determined that
          the reagent is of sufficiently high purity to permit its use without lessening the quality of the
          determination.

       7.1.1  REAGENT WATER (RW) - Purified water that does not contain any measurable
             quantity of the target analyte or interfering compounds at or above 1/3 the bromate MRL.
             The purity of the water required for this method cannot be over emphasized. The reagent
             water used  during method development was purified from tap water using a Millipore
             ELIX-3 followed by a Millipore Gradient A10 system. The water should contain no
             particles larger than 0.20 microns.

       7.1.2  ELUENT SOLUTIONS - Two hydroxide eluent concentrations were used to collect the
             data in Section 17. A potassium hydroxide eluent concentration of 10 mM was used for
             the first dimension matrix elimination separation on the AS 19 column, and 10 mM for
             the second  dimension separation on the AS24 column.  In addition, a hydroxide eluent
                                           302.0-11

-------
          concentration of 65 mM was delivered to the first dimension column after the bromate
          heart-cut step was completed to ensure the column was properly cleaned prior to the next
          analysis. As well, the second dimension AS24 column was cleaned with 65 mM
          hydroxide after the elution of bromate and equilibrated with 10 mM hydroxide prior to
          injection of the next sample. These eluents were automatically prepared using
          electrolytic eluent generation with the ICS-3000 EG Eluent Generator and EluGen
          potassium hydroxide cartridges (Sect 6.6.2). They may also be manually prepared (Sect.
          6.6.2) if adequate precautions are used to prevent carbonate formation from exposure to
          laboratory air.

   7.1.2   BROMATE  STANDARD - (BrO3~, CASRN 15541-45-4) - Ultra Scientific Cat. No.
          ULICC-010 or equivalent.

   7.1.3   SODIUM BICARBONATE - (NaHCO3, CASRN 497-19-8) - Fluka Cat. No. 71627 or
          equivalent.

   7.1.4   SODIUM BROMATE - (NaBrO3, CASRN 7789-38-0) - EM Science Cat. No. SX0385-1
          or equivalent.

   7.1.5   SODIUM CHLORIDE - (NaCl, CASRN 7647-14-5) - Fisher Scientific Cat. No. S-271
          or equivalent.

   7.1.6   SODIUM NITRATE - (NaNO3, CASRN 7631-99-4) - Fisher Scientific Cat. No. S343-
          500) or equivalent.

   7.1.7   SODIUM PHOSPHATE, DIBASIC, ANHYDROUS - (Na2HPO4, CASRN 10140-65-5)
          JT. Baker Cat. No. 4062-1 or equivalent.

   7.1.8   SODIUM SULFATE - (Na2SO4, CASRN 7757-82-6) - Fluka Cat. No.  71959 or
          equivalent.

   7.1.9   ETHYLENEDIAMINE (EDA) PRESERVATION SOLUTION (10 mg/mL) - Add 2.8
          mL of ethylenediamine (99%) (CASRN 19559-59-2) to a 250-mL volumetric flask and
          dilute to volume with reagent water.

7.2    STANDARD SOLUTIONS - When a compound purity is assayed to be 96 percent or greater,
     the weight can be used without correction to calculate the concentration of the stock standard.
      Solution concentrations listed in this section were used to develop this method and are
     included only as an example. Although estimated stability times for standard solutions are
     suggested in the following subsections, laboratories should use standard QC practices to
     determine when their standards need to be replaced.

   7.2.1   BROMATE  STANDARD SOLUTIONS - Obtain the analyte as a  solid standard or as a
          commercially prepared standard from a reputable standard manufacturer. Prepare the
          bromate stock and dilution solutions as described below.

      7.2.1.1 BROMATE STOCK STANDARD SOLUTION (SSS) (1000 mg/L BrO3') -
             Preparation of this solution is accomplished using a solid NaBrO3 standard. Weigh
                                      302.0-12

-------
             out 118.0 mg of NaBrO3 into a 100-mL volumetric flask and dilute to volume with
             reagent water. When stored in opaque, plastic storage bottles, the resulting stock
             solution may be stable for up to one year.

       7.2.1.2 BROMATE PRIMARY DILUTION SOLUTION (PDS) (10.0 mg/L BrO3") - Prepare
             the Bromate PDS by adding 1.0 mL of the Bromate SSS to a 100-mL volumetric flask
             and dilute to volume with reagent water. This solution is used to prepare the Bromate
             Secondary Dilution Standard Solution, the Bromate Fortification Solutions and the
             Calibration Solutions below. When stored in opaque, plastic storage bottles, the
             resulting solution is stable for at least one month.

       7.2.1.3 BROMATE SECONDARY DILUTION SOLUTION (SDS) (1.0 mg/L BrO3') -
             Prepare the 1.0 mg/L Bromate SDS by adding 10.0 mL of the Bromate PDS to a 100-
             mL volumetric flask and diluting to volume with reagent water. This solution is used
             to prepare the Bromate Fortification Solutions, CAL and CCC Standards listed below.
             When stored in opaque, plastic storage bottles, the resulting solution is stable for at
             least one month.

       7.2.1.4 BROMATE FORTIFICATION SOLUTIONS (BFS) (50, 200 and 500 |ig/L BrO3") -
             The Bromate Fortification Solutions are prepared by dilution of the Bromate SDS  and
             are used to fortify the Laboratory Fortified Blank (LFB), the Laboratory Fortified
             Sample Matrix (LFSM), the Laboratory Fortified Sample Matrix Duplicate (LFSMD)
             and the Laboratory Fortified Synthetic Sample Matrix  (LFSSM) with bromate. It is
             recommended that multiple concentrations be prepared  so that the fortification levels
             can be rotated or adjusted to the concentration of the target analyte in the native
             samples.  When stored in opaque, plastic storage bottles, the resulting solutions are
             stable for at least one month.

   7.2.2  LABORATORY SYNTHETIC SAMPLE MATRIX (LSSM) - Prepare a LSSM that
         contains the common anions chloride, sulfate and bicarbonate at 100 mg/L, and
         phosphate as phosphorus and nitrate as nitrogen at 10 mg/L as follows.

       7.2.2.1 Weigh out 276 mg of NaHCO3, 296 mg of Na2SO4, 330 mg of NaCl, 91.7 mg of
             Na2HPO4 and  121 mg of NaNO3.  Add these to a 2000-mL volumetric flask using a
             funnel and dilute to volume using reagent water.

7.3    CALIBRATION STANDARDS (CAL) - Prepare a calibration curve from dilutions of the
      Bromate PDS and the Bromate SDS using a minimum of five Calibration Standards, which
      span  the concentration range of interest. The lowest CAL standard must be at or below the
      MRL. An example of the dilutions used to prepare the CAL standards used to collect the data
      in Section 17 are shown in the Table below.
                                       302.0-13

-------
CALand CCC
Levels
CAL1
CAL 2
CAL 3
CAL 4
CAL 5
CAL 6
CAL 7
CAL 8
Vol. of
Br03
PDS
(HL)




30
50
100
150
Vol. of
Br03
SDS
(HL)
15
25
50
100




Final
Vol. of
Std.
(mL)
100
100
100
100
100
100
100
100
Final
Cone, of
Br03
(HS/L)
0.15
0.25
0.50
1.00
3.00
5.00
10.0
15.0
   7.4   CONTINUING CALIBRATION CHECK STANDARDS (CCC) - Prepare the CCC
         standards from dilutions of the Bromate PDS and the Bromate SDS at three concentrations
         representing the low, mid and high points of the calibration curve. The low-level CCC should
         be at or below the MRL.

   7.5   LABORATORY FORTIFIED SYNTHETIC SAMPLE MATRIX CCC STANDARD -  In
         order to ensure that the first dimension cut window is functioning properly during each
         analysis batch, a CCC is also prepared in the laboratory synthetic sample matrix.  This
         solution is termed the Laboratory Fortified Synthetic Sample Matrix (LFSSM) CCC and is
         analyzed following the last high-level CCC during each Analysis Batch.
8.  SAMPLE COLLECTION, PRESERVATION, AND STORAGE

   8.1   SAMPLE COLLECTION

      8.1.1   Grab samples must be collected in plastic or glass containers accordance with
             conventional sampling practices. 6

      8.1.2   When sampling from a cold water tap, open the tap and allow the system to flush until the
             water temperature has stabilized (usually approximately 3 to 5 minutes). Collect a
             representative sample from the flowing system using a beaker of appropriate size. Use
             this bulk sample to generate individual samples as needed. A volume of approximately
             20-mL is required for each individual sample.

      8.1.3   When sampling from an open body of water, fill a beaker with water sampled from a
             representative area.  Use this bulk sample to generate individual samples as needed.  A
             volume of approximately 20-mL of sample is required for each individual sample.

      8.1.4   Samples must be dechlorinated at the time of collection by adding EDA so that the final
             concentration in the sample container is 50 mg/L. For a 20 mL sample, this would
             require 100 uL of the EDA Preservation Solution (Sect. 7.1.9).
                                          302.0-14

-------
   8.2   SAMPLE SHIPMENT AND STORAGE - Field samples must be chilled during shipment and
         must not exceed 10 °C during the first 48 hours after collection. Field samples should be
         confirmed to be at or below 10 °C when they are received at the laboratory. Field samples
         stored in the lab must be held at or below 6 °C until analysis.

   8.3   SAMPLE HOLDING TIMES - Field samples that are collected and stored as described in
         Sections 8.1 and 8.2 may be held for up to 28 days.

9.  QUALITY CONTROL

   9.1.   Quality Control requirements include the Initial Demonstration of Capability (IDC) and
         ongoing QC requirements that must be met when preparing and analyzing field samples.  This
         section describes each QC parameter, its required frequency, and the performance criteria that
         must be met in order to meet EPA data quality  objectives. The QC criteria discussed in the
         following sections are summarized in Section 17, Tables 4 and 5. These QC requirements are
         considered the minimum acceptable QC criteria. Laboratories are encouraged to institute
         additional QC practices to meet their specific needs.

   9.2   INITIAL DEMONSTRATION OF CAPABILITY (IDC) - The IDC must be successfully
         performed prior to analyzing any field samples. Prior to conducting the IDC, the analyst must
         first meet the calibration requirements outlined in Section 10.  Requirements for the IDC  are
         described in the following sections and are summarized in Table 4.

      9.2.1  DEMONSTRATION OF LOW  SYSTEM BACKGROUND - Analyze a Laboratory
             Reagent Blank (LRB) processed through all sample collection steps outlined in Section
             8.1.  Confirm that the LRB is reasonably free of contamination (< 1/3 the MRL) and  that
             the criteria in Section 9.3.1 are met.

             NOTE: It is a good laboratory practice to include a blank in the calibration of any
             instrument. The method should also be checked for carry-over by analyzing a RW blank
             immediately following the highest CAL standard. If this RW sample does not meet the
             criteria outlined in Section 9.3.1 then carry-over is present and should be identified and
             eliminated.

      9.2.2  DEMONSTRATION OF PRECISION - Prepare and analyze 7 replicate LFBs and
             LFSSMs fortified near the midrange of the initial calibration curve. The percent relative
             standard deviation (%RSD) of the results of the replicate analyses must be < 20 percent.

                         _ Standard Deviation of Measured Concentrations
                  /o RSD —	X100
                                      Average Concentration

      9.2.3  DEMONSTRATION OF ACCURACY - Using the same set of replicate data generated
             for Section 9.2.2, calculate average recovery. The average recovery of the replicate
             values must be within ± 20 percent of the true value.

                      „._,         Average Measured Concentration  n^
                      % Recovery =	x 100
                                       Fortified Concentration

                                           302.0-15

-------
9.2.4   MINIMUM REPORTING LEVEL (MRL) CONFIRMATION - Establish a target
       concentration for the MRL based on the intended use of the method. The lowest
       calibration standard used to establish the initial calibration (as well as the low-level CCC)
       must be at or below the concentration of the MRL.  Establishing the MRL concentration
       too low may cause repeated failure of ongoing QC requirements. Confirm or validate the
       MRL following the procedure outlined below.

   9.2.4.1 Fortify and analyze seven replicate LFB at or below the proposed MRL
          concentration.  Calculate the mean (Mean) and standard deviation (S) for these
          replicates. Determine the Half Range for the Prediction Interval of Results (HRpiR)
          using the equation below.

                                      HRpm = 3.963S

           where S is the standard deviation, and 3.963 is a constant value for seven
           replicates.1

   9.2.4.2 Confirm that the upper and lower limits for the Prediction Interval of Results (PIR =
          Mean +_ HRpjR) meet the upper and lower recovery limits as shown below.

             The Upper PIR Limit must be < 150 percent recovery.

                                   Mean + HRP1R
                               FortifiedConcentration

             The Lower PIR Limit must be > 50 percent recovery.

                                  Mean -  HRPJP
                              —	——	—	x 100> 50%
                              FortijiedC oncentration

   9.2.4.3 The MRL is validated if both the Upper and Lower PIR Limits meet the criteria
          described above (Sect. 9.2.4.2).  If these criteria are not met, the MRL has been set
          too low and must be determined again at a higher concentration.

9.2.5   QUALITY CONTROL SAMPLE (QCS) - Analyze a mid-level Quality Control Sample
       (Sect. 9.3.7) to confirm the accuracy of the calibration curve fit.

9.2.6   DETECTION LIMIT DETERMINATION (DL) (optional) - While DL determination is
       not a specific requirement of this method, it may be required by various regulatory
       bodies associated with compliance monitoring. It is the responsibility of the laboratory
       to determine ifDL determination is required based upon the intended use of the data.

       Analyses for this procedure should be done over at least 3 days. Prepare and analyze at
       least 7 replicate fortified LFBs. Use the solutions described in Section 7.2.1.4 to fortify
       at a concentration estimated to be near the DL. This fortification concentration may be

                                     302.0-16

-------
          estimated by selecting a concentration at 2-5 times the noise level. Analyze the seven
          replicates through all steps of Section 11.

          NOTE: If an MRL confirmation data set meets these requirements, a DL may be
          calculated from the MRL confirmation data, and no additional analyses are necessary.

          Calculate the DL using the following equation:

                       DL = St(n- 1, 1 - alpha = 0.99)

                 where:
                       t( n -1,1 - alpha = o.99) = Student's t value for the 99% confidence level with n-1
                       degrees of freedom
                       n = number of replicates
                       S = standard deviation of replicate analyses.

          NOTE: Do not subtract blank values when performing DL calculations.

9.3     ONGOING QC REQUIREMENTS - This section describes the ongoing QC criteria that
       must be followed when processing and analyzing field samples. Table 5 summarizes these
       requirements.

   9.3.1   LABORATORY REAGENT BLANK (LRB) - A LRB is analyzed during the IDC and is
          required with each Analysis Batch (Sect. 3.1) to confirm that potential background
          contaminants are not interfering with the identification or quantitation of bromate. If
          within the retention time window of any analyte, the LRB produces a peak that would
          prevent the determination of bromate, determine the source of contamination and
          eliminate the interference before processing samples. Background from target analytes or
          contaminants that interfere with the measurement of target analytes must be < 1/3 the
          MRL.  If the target analytes are detected in the LRB at concentrations equal to or greater
          than this level, then all data for the problem analyte(s) must be considered invalid for all
          samples which yielded a positive result.

          NOTE: Although quantitative data below the MRL may not be accurate enough for data
          reporting, such data are useful in determining the magnitude of a background
          interference. Therefore, blank contamination levels may be estimated by extrapolation,
          when the concentration is below the MRL.

   9.3.2   CONTINUING CALIBRATION CHECK STANDARDS (CCC) - CCC standards are
          analyzed at the beginning of each Analysis Batch, after every ten field samples, and at the
          end of the Analysis Batch. See Section  10.3 and Table 5 for concentration requirements
          and acceptance criteria.

   9.3.3   LABORATORY FORTIFIED SYNTHETIC SAMPLE MATRIX CCC STANDARD
          (LFSSM CCC) - A CCC standard prepared in the LSSM at the same concentration as the
          high-level CCC  Standard should be analyzed at the end of each Analysis Batch. The
          LFSSM CCC is used to ensure the integrity of the sample pre-concentration/matrix
          elimination  step  and the chromatographic separation of bromate from other interfering
                                        302.0-17

-------
       anionic species in very high ionic matrices. See Section 10.3.3. and Table 5 for
       concentration requirements and acceptance criteria.

9.3.4   LABORATORY FORTIFIED BLANK (LFB) - A LFB is required with each Analysis
       Batch.  In successive analysis batches, the LFB fortification level must be rotated
       between low, medium, and high. The low concentration LFB must be at or below the
       MRL. Results of LFBs fortified at < MRL must be within 50-150% of the true value.
       Results of LFB analyses from all other concentrations must be 80-120% of the true value.
       If the LFB results do not meet these criteria, then all data for bromate must be considered
       invalid  for all samples in the Analysis Batch.

9.3.5   LABORATORY FORTIFIED SAMPLE MATRIX (LFSM) - Analysis of a LFSM (Sect.
       3.8) is required in each Analysis Batch.  The LFSM is processed and analyzed exactly
       like a traditional sample, and its purpose is to determine whether the sample matrix
       contributes bias to the analytical results.  The native concentration of the analyte in the
       sample  matrix must be determined in a separate aliquot and the measured value in the
       LFSM corrected for the native concentration.  If a variety of different sample matrices are
       analyzed regularly, for example drinking water from groundwater and surface water
       sources, performance data should be collected for each source.

   9.3.5.1 Within each Analysis Batch, a minimum of one field sample must be fortified as an
          LFSM for every 20 field  samples analyzed.  The LFSM is prepared by spiking a field
          sample with an appropriate amount of the Bromate Fortification Solutions (Sect.
          7.2. 1 .4).  The fortification should be delivered in the smallest volume possible to
          minimize dilution of the  sample. Select a fortification concentration that is  equal to
          or greater than the native concentration, if known.  Use historical data and rotate
          through the designated concentrations when selecting a fortifying concentration.

          Calculate the percent recovery (%REC) using the equation
                                         C

            A = measured concentration in the fortified sample
            B = measured concentration in the unfortified sample
            C = fortification concentration

   9.3.5.2 Recoveries for samples fortified at concentrations near or at the MRL (within a factor
          of two times the MRL concentration) should be 50-150%.  Recoveries for samples
          fortified at all other concentrations should be 80-120%. If the accuracy for any
          analyte falls  outside the designated range, and the laboratory performance for that
          analyte is shown to be in control in the CCCs and the LFSSM CCC, the recovery is
          judged to be matrix biased. The result for that analyte in the unfortified sample is
          labeled "suspect/matrix"  to inform the data user that the results are suspect due to
          matrix effects.

       9.3.5.2.1  Field samples that have a native bromate concentrations above the DL but
                below the MRL and are fortified at concentrations at or near the lowest
                                     302.0-18

-------
                calibration standard should be corrected for the native levels in order to obtain
                meaningful percent recovery values.  This example and the LRB (Sect. 9.3.1)
                are the only permitted use of analyte results below the MRL.

9.3.6  LABORATORY DUPLICATE OR LABORATORY FORTIFIED SAMPLE MATRIX
      DUPLICATE (LD or LFSMD) - Within each Analysis Batch, a minimum of one
      Laboratory Duplicate (LD) or Laboratory Fortified Sample Matrix Duplicate (LFSMD)
      must be analyzed. Laboratory Duplicates check the precision associated with laboratory
      procedures. If target analytes are not routinely observed in field samples, a LFSMD
      should be analyzed rather than a LD.

   9.3.6.1 Calculate the relative percent difference (RPD) for duplicate measurements (LDi and
          LD2) using the equation
                                  LD.-LD,
                         RPD =—!	l-	^— xlOO
                                (LD1+LD2)/2

   9.3.6.2 RPDs for Laboratory Duplicates should be < 20%. Greater variability may be
          observed when Laboratory Duplicates have analyte concentrations that are within a
          factor of 2 of the MRL. At these concentrations Laboratory Duplicates should have
          RPDs that are < 50 percent. If the RPD of any analyte falls outside the designated
          range, and the laboratory performance for that analyte is shown to be in control in the
          CCCs and LFSSM CCC, the recovery is judged to be matrix influenced. The result
          for that analyte in the unfortified field sample is labeled "suspect/matrix" to inform
          the data user that the results are suspect due to matrix effects.

   9.3.6.3 If a LFSMD is analyzed instead of a Laboratory Duplicate, calculate the relative
          percent difference (RPD) for duplicate LFSMs (LFSM and LFSMD) using the
          equation.

                                \LFSM-LFSMD
                      RPD = -^	r—xlOO
                              (LFSM+LFSMD)/2

   9.3.6.4 RPDs for duplicate LFSMs must be < 20%. Greater variability may be observed
          when LFSMs are fortified at analyte concentrations that are within a factor of 2 of the
          MRL. LFSMs fortified at these concentrations must have RPDs that are < 50%. If
          the RPD of any analyte falls outside the designated range, and the laboratory
          performance for that analyte is shown to be in control in the CCCs and LFSSM CCC,
          the precision is judged to be matrix influenced.  The result for that analyte in the
          unfortified field sample is labeled "suspect/matrix" to inform the data user that the
          results are suspect due to matrix effects.

9.3.7  QUALITY CONTROL SAMPLES (QCS) -A QCS must be evaluated as part of the IDC
      (Sect. 9.2.5) and each time new PDS solutions are prepared. If standards are prepared
      infrequently, analyze a QCS at least quarterly.  The QCS should be fortified near the
      midpoint of the calibration range and analyzed as a CCC. The acceptance criteria for the
                                    302.0-19

-------
             QCS is the same as for the mid- and high-level CCCs (Section 10.3).  If measured analyte
             concentrations are not of acceptable accuracy, check the entire analytical procedure to
             locate  and correct the problem.

   9.4    METHOD MODIFICATION QC REQUIREMENTS - The analyst is permitted to modify
          the separation technique, 1C columns, concentrator column and mobile phase composition.
          However,  any modifications must maintain the basic chromatographic elements of this new
          technique  (Sect 1.6).

       9.4.1  Each time method modifications are made, the analyst must first redetermine the cut
             window following Section 10.2.2 using bromate solutions fortified into RW and the
             LSSM (Sect. 7.2.2) and then reestablish an acceptable initial calibration (Sect.  10.2.3).
             Alternate configurations may require modification of this technique.  If alternate
             procedures are used, the analyst must still employ the LSSM and RW matrices to set the
             start and stop times of the cut windows.

       9.4.2  Repeat the procedures of the IDC (Sect. 9.2) and verify that all Ongoing QC criteria can
             be met for the proposed method modification (Sect. 9.3).

       9.4.3  The analyst is also required to evaluate method performance for the proposed method
             modifications in real matrices that span the range of waters that the laboratory  analyzes.
             This additional step is required because modifications that perform acceptably  during the
             IDC, which is conducted in reagent water and the LSSM, can fail ongoing method QC
             requirements in real matrices due to common method interferences. If, for example, the
             laboratory analyzes finished waters from both surface and groundwater municipalities,
             this requirement can be accomplished by assessing precision and accuracy (Sects. 9.2.2
             and 9.2.3) in a surface water with moderate to high total organic carbon ( e.g., 2 mg/L or
             greater) and a hard groundwater (e.g., 250 mg/L or greater).

       9.4.4  The results of Sections 9.4.1 - 9.4.3 must be appropriately documented by the  analyst and
             should be independently assessed by the laboratory's QA officer prior to its application to
             field samples.

          9.4.4.1 When implementing method modifications, it is the responsibility of the laboratory to
                 closely review the results of Ongoing QC, and in particular, the results associated
                 with the LFSM (Sect. 9.3.5), the LFSMD (Sect. 9.3.6) and the LFSSM CCCs (Sect.
                 9.3.3).  If repeated failures are noted, the modification must be abandoned.

10. CALIBRATION AND STANDARDIZATION

   10.1   Demonstration and documentation of acceptable initial calibration for bromate is required
          prior to conducting the IDC and before any field samples are analyzed. If the initial
          calibration is successful, continuing calibration check standards are  required at the beginning
          and end of each Analysis Batch, as well as after every tenth field sample.

   10.2   INITIAL CALIBRATION - The initial calibration must be established prior to conducting
          the IDC (Section 9.2) and may be reestablished prior to analyzing field samples. However, it
          is permissible to verify the calibration with daily  CCCs.  Calibration should be performed
                                            302.0-20

-------
   using peak areas and the external standard technique.  Calibration using peak heights is not
   permitted.

10.2.1 INSTRUMENT CONDITIONS - Establish proper operating conditions. Operating
      conditions used during method development are described in Section 17 Table 1.
      Conditions different from those described may be used if the IDC QC criteria in Section
      9.2 are met. The procedure used to establish the first dimension cut window is provided
      in Section 10.2.2 below.

      NOTE: For 2D-IC techniques, increased method sensitivity is achieved by injecting
      larger volumes.  Increased injection volume must be carefully evaluated for each set of
      instrument conditions. For the conditions and equipment reported in this method, the
      maximum recommended injection volume is  1.0 mL. This is because bromate elutes just
      prior to chloride on the first dimension column. Injection volumes above 1.0 mL may
      cause potential chloride interference.

10.2.2 FIRST DIMENSION CUT WINDOW DETERMINATION - Since a large volume (1.0
      mL) is injected onto the first dimension column, the affect of high ionic strength matrices
      on the bromate retention time requires careful consideration.  Determination of the cut
      window is the first step in separating bromate from other interfering anionic matrix
      species. Setting of the cut window in the first dimension must include evaluation of the
      retention time for bromate in both RW and the LSSM.

      NOTE: The procedure described below is applicable to the equipment configuration
      listed in Section 6.  However, since acceptable modifications to this method require
      heart-cutting of bromate, the trapping of bromate on a concentrator column, and the
      subsequent separation and quantitation on a dissimilar analytical  column, all modified
      versions of this procedure require the proper determination of the cut window. Similar
      techniques must therefore be developed and used by analysts using modified methods.

   10.2.2.1   DETERMINING THE START TIME OF THE CUT WINDOW - Inject an
          aliquot (1.0 mL) of a  15 ug/L bromate fortification in the LSSM (without injection
          valve #2 on system #2 being activated) to determine the start time for the cut window.
          The start time for the cut window in the first dimension should be set at 0.50 minutes
          prior to the start of the bromate elution in the LSSM.

          NOTE: The bromate in the LSSM may not appear as a distinct peak, but rather as a
          broad, smeared peak on the first dimension column.  However, a distinct rise in
          baseline is  evident when bromate starts to elute from the column (see Figure 3).

   10.2.2.2   DETERMINING THE STOP TIME OF THE CUT WINDOW - The stop time
          for the cut window is established using reagent water.  Inject an aliquot (1.0 mL) of a
          15 ug/L bromate standard in RW and determine when bromate is completely eluted
          off the first dimension column. The stop time for the cut window in the first
          dimension  should be set at 0.20 minutes after the bromate peak in RW returns to
          baseline.
                                    302.0-21

-------
             NOTE:  Since the stop time for the cut window is also the time at which injection
             valve #2 on system #2 is switched to the inject position, the elution of bromate must
             be completed at least 15 seconds before injection valve #2 is switched into the inject
             position when this instrument configuration is used (see Figure 3).

   10.2.3  CALIBRATION - Prepare a set of at least five CAL standards as described in Section
          7.3. The lowest concentration CAL standard must be at or below the MRL.  TheMRL
          must be confirmed using the procedure outlined in Section 9.2.4, after establishing the
          initial calibration. Calibrate the conductivity detector using the external standard tech-
          nique.  Calibration curves may be generated using the 1C data system through the use of
          first (linear) or second (quadratic) order calibration curves.   A quadratic fit is
          recommended for this method. Concentration-based weighting may also be used;
          however, if this approach is chosen the analyst must confirm this fit does not introduce
          bias in the higher concentration region of the curve.

       10.2.3.1    CALIBRATION ACCEPTANCE CRITERIA - The validation of the calibration
             is determined by calculating the concentration of the analyte from each of the
             analyses used to generate the calibration curve. At least one CAL standard must be at
             or below the MRL.  This calibration point should calculate to be 50 to 150 percent of
             its true value. All other higher concentration calibration points should calculate to be
             80 to 120 percent of their true values. If these criteria cannot be met, the analyst will
             have difficulty meeting ongoing QC criteria. In this case, corrective action should be
             taken to reanalyze the calibration standards, restrict the range of calibration, or select
             an alternate method of calibration.

10.3    CONTINUING CALIBRATION CHECK (CCC) STANDARDS - The CCCs verify the
       calibration at the beginning and end of each group of analyses, as well as after every 10th
       field sample. The LRBs, LFBs, LFSSMs, LFSMs, LFSMDs, and CCCs are not counted as
       field samples. The beginning CCC for each  Analysis Batch must be at  or below the MRL in
       order to verify instrument sensitivity and the accuracy of the calibration curve prior to the
       analysis of any field samples. Subsequent CCCs should  alternate between a medium and
       high concentration.

   10.3.1  Inject an aliquot of the CCC standards and analyze with the same conditions used during
          the initial calibration.

   10.3.2  Calculate the concentration of the analyte in the CCC standards. The calculated amount
          for the analyte  for mid and high level CCCs must be within ± 20 percent of the true
          value. The calculated analyte amount for the lowest CCC level must be within ± 50
          percent of the true value.  If these conditions do not exist, then all data for the analyte
          must be considered invalid, and remedial action (Sect.  10.3.4) must be taken which may
          require recalibration.  Any results from field samples that have been analyzed since the
          last acceptable calibration verification are invalid.

       10.3.2.1    The analyst should carefully review all first-dimension chromatograms for the
             CCCs to ensure that the entire chromatographic peak elutes within the cut window. If
             this is not the case, the analyst should redetermine the cut window as  per Section
             10.2.2.
                                        302.0-22

-------
       10.3.3 LABORATORY FORTIFIED SYNTHETIC SAMPLE MATRIX CCC STANDARD
             (LFSSM CCC) - As chromatographic columns age, column performance can deteriorate.
             This deterioration will typically result in a decreased retention time for bromate, which
             can affect the bromate retention time in high inorganic matrices.  A LFSSM CCC must be
             analyzed daily in each Analysis Batch at a concentration equal to the high-level CCC to
             ensure that the first dimension cut window is functioning properly for high inorganic
             strength waters. The QC acceptance criteria for the LFSSM CCC is the same as for the
             high-level CCCs (within ± 20%). If these conditions do not exist, then all data for the
             analyte must be considered invalid, and remedial action should be taken which may
             require the reestablishing the cut window (Sect.  10.2.2) and recalibration.

          10.3.3.1   The analyst should carefully review all first-dimension chromatograms for the
                LFSSM CCC to ensure that the entire chromatographic peak elutes within the cut
                window. If this is not the case, the analyst should redetermine the cut window as per
                 Section 10.2.2.

       10.3.4 REMEDIAL ACTION - Failure to meet CCC or LFSSM CCC QC performance criteria
             requires remedial action. Maintenance such as confirming the integrity of the trapping
             efficiency of the concentrator column and matrix elimination step and/or regenerating or
             replacing the 1C guard and analytical columns require returning to the initial calibration
             step (Sect. 10.2).

11.  PROCEDURE

   11.1   Important aspects of this  analytical procedure include proper field sample collection,
          preservation and storage (Sect. 8), ensuring that the  instrument is properly calibrated (Sect.
          10.2) and that all required QC are met (Sect. 9) during each Analysis Batch. This section
          describes the procedures for field sample preparation and analysis. If alternative
          instrumentation and/or columns to those listed in this method are used, the procedure
          outlined in Sections 9.2 must be followed prior to analyzing field samples.

   11.2   SAMPLE PREPARATION

       11.2.1 Collect and store field samples as described in Section 8.

   11.3   SAMPLE ANALYSIS

       11.3.1 Establish the instrument operating conditions as described in Table 1 of Section 17.
             Confirm that the analyte retention times for the calibration standards are stable.

       11.3.2 Establish a valid initial calibration following the procedures outlined in Section 10.2 or
             confirm that the calibration is still valid by running a low-level CCC as described in
             Section 10.3. If establishing an initial calibration for the first time, complete the IDC as
             described in Section 9.2.
                                            302.0-23

-------
11.3.3 Analyze field and QC samples at their required frequencies using the same conditions
      used to collect the initial calibration. Table 6 shows an acceptable analytical sequence
      that contains all method-required QC samples.

11.3.4 COMPOUND IDENTIFICATION - Establish an appropriate retention time window for
      bromate elution from the second dimension column in order to identify it in QC and field
      sample chromatograms.

   11.3.4.1    Since the ionic strength of drinking water matrices may vary considerably, the
              cut window (Sect. 10.2.2) for bromate in the first dimension must be set wide
              enough to account for the variability in the  ionic strength of the drinking
             water matrices.  A window of approximately 2 minutes (using a 1.0 mL injection
             volume) was been found to be acceptable for the system used during method
              development. If the cut window is not set properly, analyte recoveries may be
             reduced without affecting the second-dimension retention times.

   11.3.4.2   Retention times in the second dimension should be very stable as long as the first-
              dimension heart-cut window has been set properly (Sect. 10.2.2). This is because
              a majority of the common anions, which alter the ionic strength of field samples
              causing sample-to-sample variation in 1C retention time, are eliminated during the
             heart-cut procedure. Retention times measured for bromate in RW and for the
             LFSSM were essentially identical during method development.

   11.3.4.3    The QC requirements for each analysis batch include a first-dimension printout of
             the final CCC and the high level LFSSM CCC chromatograms to ensure the
             heart-cut window is  functioning properly. The analyst should also review all
             first-dimension chromatograms to ensure acceptable chromatographic
             performance within the first-dimension cut window. For example, if a sample had
              a much higher conductivity (significantly above the LFSSM) caused by either
             very high levels of interfering anions or bromate, the first-dimension cut window
             would be overwhelmed.  In such an instance, the sample should be diluted and
             reanalyzed.

   11.3.4.4   High ionic strength matrices have the potential to cause an increase in background
              conductivity and severe tailing as the other anions elute from the first dimension
              column and cause the bromate retention time to decrease.

11.3.5 EXCEEDING CALIBRATION RANGE - The analyst must not extrapolate beyond the
      established calibration range. If an analyte result exceeds the range of the initial
      calibration curve, the field sample may be diluted with reagent water and the diluted field
      sample re-injected.  Incorporate the dilution factor into final concentration calculations.
      The dilution will also affect the bromate MRL.
                                     302.0-24

-------
12. DATA ANALYSIS AND CALCULATIONS

   12.1  Identify the analyte present in the field and QC Samples as described in Section 11.3.4.

   12.2  Calculate the bromate concentrations using the multi-point calibration established in Section
         10.2.  Quantify only those values that fall between the MRL and the highest calibration
         standard. Field samples with target analyte responses that exceed the highest calibration
         standard require dilution and reanalysis (Sect. 11.3.5).

       12.2.1 As noted in Section 9.3.5.2.1, it may be necessary to extrapolate below the MRL to
             estimate contaminants in LRBs and to correct for native levels of bromate below the
             MRL when field samples are fortified at or near the MRL. These are the only permitted
             use of analyte results below the MRL.

   12.3  Calculations must utilize all  available digits of precision, but final reported concentrations
         should be rounded to an appropriate number of significant figures (one digit of uncertainty),
         typically two, and not more than three significant figures.

   12.4  Prior to reporting data, the laboratory is responsible for assuring that QC requirements have
         been met or that any appropriate qualifier is documented.

13. METHOD PERFORMANCE

   13.1  PRECISION, ACCURACY AND DETECTION LIMITS - Tables for these data are
         presented in Section 17. The LCMRL for bromate is presented in Table 2 and was calculated
         using a procedure described elsewhere.1   The DL data is also reported in Table 2. Single
         laboratory precision and accuracy data are presented in Table 3.

   13.2  Figure 3 is a representation of both the first and second dimension chromatograms for a 15
         ug/L bromate fortification in the LSSM and Figure 4 shows similar chromatograms for a 0.50
         Hg/L bromate fortification to a municipal ground water, disinfected with chlorine.

14. POLLUTION PREVENTION

   14.1  For information about pollution prevention that may be applicable to laboratory operations,
         consult "Less is Better: Laboratory Chemical Management for Waste Reduction" available
         from the American Chemical Society's Department of Government Relations and Science
         Policy, 1155 16th Street N.W., Washington, D.C., 20036, or on-line at:
         http://www.ups.edu/community/storeroom/Chemical Wastes/wastearticles.htm.

15. WASTE MANAGEMENT

   15.1  The analytical procedures described in this method generate relatively small amounts of waste
         since only small amounts of reagents are used. The matrices of concern are finished drinking
         water or source water.  However, the Agency requires that laboratory waste management
         practices be conducted consistent with all applicable rules and regulations, and that
         laboratories protect the air, water, and land by minimizing and controlling all  releases from
                                            302.0-25

-------
         fume hoods and bench operations. Also, compliance is required with any sewage discharge
         permits and regulations, particularly the hazardous waste identification rules and land disposal
         restrictions. For further information on waste management, see the publications of the
         American Chemical Society's Laboratory Environment, Health & Safety Task Force on the
         Internet at http://membership.acs.Org/c/ccs/publications.htm.  Or see "Laboratory Waste
         Minimization and Pollution Prevention," Copyright ©  1996 Battelle Seattle Research Center,
         which can be found at http://www.p2pays.org/ref/01 /text/00779/index2.htm.

16. REFERENCES

   1.  Winslow, S. D.; Pepich, B. V.; Martin, J. I; Hallberg, G.  R.; Munch, D. I; Frebis, C. P.;
       Hedrick, E. J.; Krop, R. A. Statistical Procedures for Determination and Verification of
       Minimum Reporting Levels for Drinking Water Methods. Environ. Sci. & Technol. 2006,  40,
       281-288.

   2.  Glaser, J.A.;  Foerst, D.L.; McKee, G. D.; Quave, S. A.; Budde, W. L. Trace Analyses for
       Wastewaters, Environ. Sci. & Technol. 1981, 15, 1426-1435.

   3.  Safety in Academic Chemistry Laboratories; American Chemical Society Publication,
       Committee on Chemical Safety, 7th Edition: Washington,  D.C., 2003.

   4.  Occupational Exposures to Hazardous Chemical in Laboratories; 29 CFR 1910.1450,
       Occupational  Safety and Health Administration,  1990.

   5.  "Guidelines, Recommendations, and Regulations for Handling Antineoplastic Agents", Center
       for Disease Control, National Institute for Occupational Safety and Health,
       http://www.cdc.gov/niosh/topics/antineoplastic/pubs.htmltfb

   6.  Standard Practice for Sampling Water from Closed Conduits; ASTM Annual Book of Standards,
       Section 11, Volume 11.01, D3370-82, American Society for Testing and Materials: Philadelphia,
       PA, 2008.
                                            302.0-26

-------
 17.0 TABLES, DIAGRAMS, FLOWCHARTS, AND VALIDATION DATA
 TABLE 1. INSTRUMENTAL CONDITIONS USED DURING METHOD DEVELOPMENT

Standard Conditions and Equipment for First Dimension Analyses a'b:
Ion Chromatograph:
Sample loop:
Load Volume:
Eluent Generator:
Eluent Flow:
Autosampler:

Columns :

Typical System Back-pressure:
Conductivity Suppressor:

Chromatography Module
Detector:
Total analysis time:
Dionex ICS-3000 Dual System
1.0 mL
1000 |iL
Dionex ICS-3000 EG Eluent Generator Module (P/N 061714), with
dual channel EluGen Cartridges (P/N 058900), isocratic 10 mM
potassium hydroxide step changed or ramped to 65 mM.potassium
hydroxide following the elution of bromate (when injection valve #2 is
switched to the inject position)
1.0 mL/min
AS Autosampler Module (P/N 063105) Sample Prep options and a
large volume (8.2 mL) sample needle assembly (P/N 061267)
Dionex lonPac AG 19 Guard 4 x 50 mm (P/N 062887)
Dionex lonPac AS 19 Analytical 4 x 250 mm (P/N 062885)
-2300 psi
Dionex ASRS Ultra II, 4-mm (P/N 061561) with 4-mm Carbonate
Removal Device (CRD, P/N 062983)
DC-Module (P/N 061793) maintained @ 30° C
Conductivity Detector with integrated cell (P/N 061716) held @ 35° C
35 min
Standard Conditions and Equipment for Second Dimension Analyses a'b:
Ion Chromatograph:
Sample loop:
Load Volume:

Eluent Generator:
Eluent Flow:
Columns :

Typical System Back-pressure:
Conductivity Suppressor:

Chromatography Module
Detector:
Total analysis time:
Dionex ICS-3000 Dual System
Dionex UTAC-ULP1, 5 x 23-mm, concentrator column (P/N 063475)
Cut-window time set per procedures in Section 10.2.2 (approximately
2.0 mL)
Dionex ICS-3000 EG Eluent Generator Module (P/N 061714), with
dual channel EluGen Catridges (P/N 058900), isocratic 10 mM
potassium hydroxide step changed or ramped to 65 mM potassium
hydroxide following the elution of bromate for approximately 10
minutes and re-equilibrate at 10 mM hydroxide prior to injection
0.25 mL/min
Dionex lonPac AG 24 Guard 2 x 50 mm (P/N 064151)
Dionex lonPac AS 24 Analytical 2 x 250 mm (P/N 064153)
-2300 psi
Dionex ASRS Ultra II, 2-mm (P/N 061562) with 2-mm Carbonate
Removal  Device (CRD, P/N 062986)
DC-Module (P/N 061793) maintained @ 30° C
Conductivity Detector with integrated cell (P/N 061716) held @ 35° C
35 min
   (a) Mention of trade names or commercial products does not constitute endorsement or
      recommendation for use. (b) It is the responsibility of the analyst to assure that the instruments,
      supplies, and conditions used are capable of meeting all method QC requirements.
                                          302.0-27

-------
 TABLE 2. LOWEST CONCENTRATION MRL AND DL FOR BROMATE
           WITH 2-D 1C
Injection
Volume
1.0 mL
Analyte
BrO3"
LCMRLa (^ig/L)
0.18
*DL (us/L)
0.12b
       aLCMRL was calculated according to the procedure in reference 1.
       *The DL was calculated from data acquired on a single day.
        Replicate fortifications at 0.20 |ig/L.
TABLE 3. 1C PRECISION AND RECOVERY DATA FOR BROMATE IN VARIOUS
          MATRICES WITH 2-D 1C (n=7) USING A 1.0-mL INJECTION VOLUME
Matrix
Reagent Water

*LFSSM

Ground water

Surface water 1

Spiked
Cone.
(UR/L)
0.50
5.0
0.50
5.0
0.50
5.0
0.50
5.0
Unspiked
Cone.
(UR/L)
<0.18
<0.18
<0.18
<0.18
0.63
0.63
0.46
0.46
Mean
% Rec.
101
99.1
104
96.2
96.1
100
89.8
103
% RSD
1.1
1.2
2.6
1.0
1.7
0.94
0.84
1.2
                   *Described in Section 3.11 and 7.2.2
                                         302.0-28

-------
TABLE 4.  INITIAL DEMONSTRATION OF CAPABILITY QUALITY CONTROL
           REQUIREMENTS
   Method
  Reference
   Requirement
    Specification and
       Frequency
    Acceptance Criteria
  Section
  9.2.1
Demonstration of
Low System
Background
Analyze a LRB prior to
any other IDC steps.
Demonstrate that bromate is
below 1/3 of the MRL (Section
9.3.1) and that possible
interferences from sampling
protocols do not prevent the
identification and
quantification of bromate.
  Section
  9.2.2
Demonstration of
Precision
Analyze 7 replicate LFBs
and LFSSMs fortified near
the mid-point of the
calibration curve
%RSDmustbe < 20%.
  Section
  9.2.3
Demonstration of
Accuracy
Calculate average recovery
for replicates used in
Section 9.2.2.
Mean recovery within + 20%
of true value.
  Section
  9.2.5
Quality Control
Sample
During IDC, each time a
new analyte PDS is made,
every time the instrument
is calibrated and at least
quarterly.	
The result for bromate must be
within 80-120% of the true
value.
                                          302.0-29

-------
TABLE 5.   ONGOING QUALITY CONTROL REQUIREMENTS (SUMMARY)
   Method
  Reference
  Requirement
  Specification and Frequency
    Acceptance Criteria
  Section 8.3
Sample Holding
Time
28 days when collected and stored
according to sections 8.1 and 8.2
with appropriate storage.	
Sample results are valid only if
samples are analyzed within
sample hold time.	
  Section
  9.3.1
Laboratory
Reagent Blank
(LRB)
Analyze a LRB as part of the IDC
(Section 9.2), as part of each
calibration, and with each
Analysis Batch
Demonstrate that bromate is
below 1/3 of the MRL and that
possible interferences from
sampling protocols do not
prevent the identification and
quantification of bromate.	
  Section
  9.3.2
  And 10.3
Continuing
Calibration Check
(CCC) Standards
Verify initial calibration by
analyzing a low-level CCC at the
beginning of each Analysis Batch.
Subsequent CCCs are required
after every 10 field samples, and
after the last field sample in a
batch.
Low CCC - at or below the MRL
concentration
Mid CCC - near midpoint in
calibration curve
High CCC - near the highest
calibration standard.
For each CCC the result must
be:
For CCCs < MRL:
% Rec within + 50% of the true
value

For CCCs > MRL: %Rec
within + 20% of the true value

Recalibration is recommended
if these criteria are not met.
  Section
  9.3.3
Laboratory
Fortified
Synthetic Sample
Matrix CCCs
(LFSSM CCC)
In order monitor the cut window
during an Analysis Batch, a high-
level CCC standard, prepared in
the LFSSM (Sect. 9.3.3) is also
required near the beginning of
each Analysis Batch.	
For the LFSSM CCC the result
must be within + 20% of the
true value
  Section
  9.3.4
Laboratory
Fortified Blank
(LFB)
Analyze a LFB with each analysis
batch, rotating between low,
medium and high concentration
from batch to batch.
                                Low LFB - at or below the MRL
                                concentration
                                Mid LFB - near midpoint in
                                calibration curve
                                High LFB - near the highest
                                calibration standard.
For each LFB the result must
be:
For LFBs < MRL:
% Rec within + 50% of the true
value

For LFBs > MRL: %Rec
within + 20% of the true value

Recalibration is recommended
if these criteria are not met.
                                           302.0-30

-------
TABLE 5. ONGOING QUALITY CONTROL REQUIREMENTS (Continued)
   Method
  Reference
   Requirement
     Specification and
        Frequency
     Acceptance Criteria
  Section
  9.3.5
Laboratory
Fortified Sample
Matrix (LFSM)
Analyze one LFSM per
Analysis Batch (20 field
samples or less).  Fortify the
LFSM with bromate at a
concentration close to but
greater than the native
concentration (if known).
Calculate LFSM recoveries.
Recoveries for the LFSM must
be calculated (Sect. 9.3.5.2).
The result must be:

For LFSMs < MRL: % Rec
within + 50% of the true value

For LFSMs > MRL: %Rec
within + 20% of the true value
  Section
  9.3.6
Laboratory
Duplicate (LD) or
Laboratory
Fortified Sample
Matrix Duplicate
(LFSMD)
Analyze at least one LD or
LFSMD daily, or with each
Analysis Batch (20 samples
or less), whichever is more
frequent.
RPD must be calculated (Sect.
9.3.6.1 for LD and Sect. 9.3.6.3
for LFSMD). The result must
be:
For LFSMDs < MRL: % Rec
within + 50% of the true value

For LFSMDs > MRL: %Rec
within + 20% of the true value
  Section
  9.3.7
Quality Control
Sample (QCS)
During IDC, each time a new
analyte PDS is made, every
time the instrument is
calibrated and at least
quarterly.
Results must be within + 20%
of the expected value.
  Section
  10.2
Initial Calibration
Use external standard
calibration technique to
generate a first or second
order calibration curve.  Use
at least 5 standard
concentrations.
Check the calibration curve
as described in Section 10.2.

Analyze a QCS near the mid-
point of the calibration
curve.
When each calibration standard
is calculated as an unknown
using the calibration curve, the
result should be:
For CALs < MRL: % Rec
within + 50% of the true value

For CALs > MRL: %Rec
within + 20% of the true value

Recalibration is recommended
if these criteria are not met.
                                           302.0-31

-------
TABLE 6. SAMPLE ANALYSIS BATCH WITH QC REQUIREMENTS
Injection
#
1
2
O
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
Sample
Description
Laboratory Reagent Blank (LRB)
CCCattheMRL
LFSSM CCC at high level
LFB
Sample 1
Sample 2
Sample 2 - Laboratory Fortified Sample Matrix (LFSM)
Sample 2 - Laboratory Fortified Sample Matrix Duplicate
(LFSMD)
Sample 3
Sample 4
Sample 5
Sample 6
Sample 7
Sample 8
Sample 9
Sample 10
CCC at mid level
Sample 1 1
Sample 12
Sample 13
Acceptance
Criteria
< 1/3 MRL
Recovery of 50- 150%
Recovery of 80 - 120%
< MRL 50-150% of value
> MRL 80-120% of value
normal analysis
normal analysis
Recovery of 80- 120%
%RPD = ±20%
normal analysis
normal analysis
normal analysis
normal analysis
normal analysis
normal analysis
normal analysis
normal analysis
Recovery of 80 - 120%
normal analysis
normal analysis
normal analysis
                         CONTINUED on NEXT PAGE
                                  302.0-32

-------
TABLE 6. (Continued)
Injection
#
21
22
23
24
25
26
27
28
Sample
Description
Sample 14
Sample 15
Sample 16
Sample 17
Sample 18
Sample 19
Sample 20
CCC at high level*
Acceptance
Criteria
normal analysis
normal analysis
normal analysis
normal analysis
normal analysis
normal analysis
normal analysis
Recovery of 80 - 120%
* also requires first dimension chromatogram printout
                                          302.0-33

-------
          Figure 1. EPA Method 302.0 System Schematic
  1st Dimension
             2nd Dimension
waste	"

  Large Loop
   	Load
   — Inject
                    Autosampler
Injection Valve 1
 4mmAG/AS19
   Column 1
  External Water
  Suppressor 1
                  CRD1
             External Water- •
                                          waste —
                                               I	
                               Injection Valve 2
  CD 2

   CRD2

^Suppressor 2
                                    2mm AG/AS24 Column 2
   Concentrator
     Column
    	Transfer to 2-D
    — Load Concentrator
                                              I    i	1
                                            waste
                            302.0-34

-------
Figure 2. Setting the start and stop time for the cut window in the first dimension using a 1.0-mL
injection volume of a 15 ug/L BrOs" fortification in the LSSM and RW.
      0.60
      0.50
      0.40-
      0.30
           MS
            Set start of cut window at 8.0 min
      0.20
      0.10-
            Return to baseline in RW@ 9.8 min
            Start of bromate peak in LSSM @
            8.5 min
     -0.10
                                                                        Set end of cut window at
                                                                        10.0 min
         0.0    1.0
3.0
5.0
7.0
9.0
11.0
13.0
                                              302.0-35

-------
Figure 3.  First and Second Dimension Chromatogram of 15.0 ug/L bromate fortification in the
LSSM.
  0.40 -
  -0.10-
-0 15 n ' ' ' i '  ' ' i ' ' ' i ' '  ' i ' ' ' i
    0.0       4.0       8.0
                                                                                     mm
                                 12.0
16.0
20.0
24.0
28.0
32.0
37.0
                                            302.0-36

-------
Figure 4.  Typical 1st and 2nd Dimension Chromatogram of a municipal groundwater disinfected
with chlorine and fortified with 0.50 ug/L bromate.  The native level of bromate was 0.41 ug/L and
recovery was 104 percent.
 0.250
 0.200
 0.150-
 0.100
 0.050-
 0.000
 0.050
 -0.100] . .
     0.0
        MS
4.0
         8.0
12.0
16.0
20.0
24.0
28.0
32.0
mm

   37.0
                                            302.0-37

-------