Toxics in Schools, Hexavalent Chromium SOP
                                        July 30, 2009
    STANDARD OPERATING PROCEDURE
                    FOR
MEASUREMENT OF HEXAVALENT CHROMIUM
                 USING THE
    BGI PQ167R LOW VOLUME SAMPLER
          U.S. Environmental Protection Agency
      Region 4, Science and Ecosystem Support Division
               Athens, Georgia, 30605
                   Page 1 of 19

-------
                                 Toxics in Schools, Hexavalent Chromium SOP
                                                          July 30, 2009
                          Acknowledgement

This Standard Operating Procedure (SOP) was developed by EPA Region 4,
Science and Ecosystem Support Division. This SOP is based on the
Commonwealth of Kentucky's ambient monitoring SOP template. Special thanks
to BGI Inc. and ERG for operational content and illustrations.

For questions or comments please contact:

Greg Noah, EPA, Region 4, SESD at noah.greg@epa.gov or 706-355-8635
Mike Jones, EPA-OAQPS-AQAD at iones.mike(S)epa.gov or 919-541-0528
                            Page 2 of 19

-------
                                     Toxics in Schools, Hexavalent Chromium SOP
                                                                July 30, 2009
                             Table of Contents

I.  Introduction	4
II.  Installation	7
      A. Sampler Siting	7
      B. Sampler Installation	7
            Cr+6 Retrofit Instructions	7
III. Operating Procedure	9
      A. Equipment and Supplies	9
      B. Sampler and Sample Media Receipt Activities	9
      C. Verification	10
      D. Calibration	11
      E. Conducting the Sampling Event	12
            Site Arrival Daily Activities	12
            Preparing Sampler for a Sampling Event	13
            Installing the ERG Cr+6 Filter Holder Module	16
            Sample  Recovery and Data Collection	16
            Sample  Shipping	17
IV. Quality Assurance	18
      A. Flow Calibration	18
      B. Flow Verifications	18
      C. Independent Audits	18
V.  Data Forms	18
      ERG Ambient Hexavalent Chromium Data Sheet	19
                                Page3 of 19

-------
                               Toxics in Schools, Hexavalent Chromium SOP
                                                          July 30, 2009
INTRODUCTION

This procedure is designed to provide instruction on collecting hexavalent
chromium (Cr+6) in air using the BGI PQ167R air sampler for metals analysis.

The BGI PQ100 is an "Intelligent Air Pump" that can monitor its own airflow
rate and thereby adjust the pump speed to compensate for changes in load
pressure and/or other forces which would otherwise hamper the flow of air
through a filter (or sample collector). The PQ100 unit can be programmed to
begin its sampling job at a specific date, time, and stop sampling after the user
defined run time is depleted. However, the sampling time should always be
24 hours for Cr+6 sampling the Toxics in Schools Study.

The PQ100 was designed to operate from  1 standard liter per minute (1000 cc
per minute) to 25.0 standard liters per minute and is unaffected by changes in
ambient temperature and barometric pressure. The flow  rate precision is
guaranteed to 2% of the calibration set point.

This SOP is designed to be a step by step method for operating the sampler
to be used in conjunction with the manufacturer's operators manual.
Laboratory Analysis Methodology may be referenced by  contacting the
Eastern Research Group (ERG) directly at 919-468-7800 or by email
Julie.Swift@erg.com.  Maintenance and troubleshooting  should be conducted
using the BGI167R operator's manual.
                          Page 4 of 19

-------
                           Toxics in Schools, Hexavalent Chromium SOP
                                                        July 30, 2009
    FIGURE 1. Schematic of PQ167 Sampling System
  (Cr+6 filter holder apparatus replaces PM10 inlet head)
  SIZE SELECTIVE
  INLET OR FILTER
           MASS FLOW
            SENSOR
PULSATION DAMPING
    VOLUME
                        PUMP
                       MICRO PROCESSOR
VISUAL DATA DISPLAY
                                 BATTERY
                                                 BATTERY CHARGER
                                                 A.C. POWER SUPPLY
                      Page 5 of 19

-------
                                   Toxics in Schools, Hexavalent Chromium SOP
                                                             July 30, 2009
                 FIGURE 2. PQ167R with Mounting Stand
(Cr+6 filter holder apparatus replaces PM10 inlet head and filter cassette module
                and downtube assembly brace are not used)
                              Page 6 of 19

-------
                                Toxics in Schools, Hexavalent Chromium SOP
                                                           July 30, 2009
INSTALLATION

A.   Sampler Siting

     Check the areas for safety.  Ensure there will be enough room for the
     operator to move freely while working, and ensure physical conditions of
     the location will allow the operator to work safely.

     The sampler should be set in a location unobstructed from any side.  No
     tree limbs or other hanging obstructions should be above the sampler.  It
     is suggested that the horizontal distance from the sampler to the closest
     vertical obstruction higher than the sampler be at least twice the height of
     the vertical obstruction.  There should be no sources located nearby that
     may bias sampling measurements.

     Locate the sampler on a reasonably level structure at a height between
     two (2) and fifteen (15) meters above the ground.

B.   Sampler Installation

     Assemble the sampler according to Figure 3 below omitting the
     installation of the PM10 inlet head, filter cassette holder assembly, and
     downtube assembly brace.  For detail and illustration,  refer to the BGI
     PQ167 Quick Start document, pages 2 through 8.

     Cr+6 Retrofit Instructions

     1.    The sampling unit, at this point, should have legs mounted on the
          stand, and the pump and power components should be secured in
          the stand according to the PQ167 Quick Start document.  The
          downtube, PM10  inlet head, and filter cassette holder assembly
          should NOT be installed.

     2.    Install the downtube on the top of the cylindrical  mount on the
          stand.  The mount should have tubing leading from the port on  its
          side to the inlet on the pump module.

     3.    The total ERG Cr+6 filter holder apparatus consists of a BGI flow
          adapter with shut-off valve, stainless steel connector fitting, a
          length of "U" shaped stainless steel % inch tubing, ERG filter
          assembly, and a glass funnel. The ERG filter assembly and
          glass funnel will be provided for each sampling run and
          should not be installed until a sampling run is setup.
          Place this apparatus (without ERG filter assembly) on  the top
          of the downtube, and ensure that the shut-off valve is in the
          open position.
                           Page 7 of 19

-------
        4.
                        Toxics in Schools, Hexavalent Chromium SOP
                                                     July 30, 2009

The open end of the stainless steel tubing should be capped when
sampling is not in progress to prevent contamination.
                    Figure 3.  Sampler Assembly Diagram
(Cr+6 filter holder apparatus replaces PM10 inlet head and filter cassette module and downtube
                          assembly brace are not used)
                       01     PM10 Inlet head
                       03     Water collection bottle
                       11A    Tripod frame
                       12     Rubber hose
                       34     Hose adapter
                       160    Sampler leg
                       161    Filter cassette holder assembly
                       162    Downtube
                       163    Downtube assembly brace
                                Page 8 of 19

-------
                                Toxics in Schools, Hexavalent Chromium SOP
                                                           July 30, 2009
     Place and level the sampler on site.  To secure the sampler and protect
     membrane roofs, 2x4 wooden studs may be cut into one foot sections
     and fastened to the feet of the legs using lag bolts.  Place sand bags on
     these skids to prevent tipping of the sampler.

     Connect the sampler to a grounded electrical outlet with 115 volts, and at
     least 5 amp service.  Protect the connector from precipitation by fastening
     beneath the sampler or wrapping it with plastic tape.

     If operating using a deep cycle marine battery for power, install the
     external power cord by screwing the round harness into the "utility
     adapter" port on top of the sampler. Attach the positive and negative
     contacts to the deep cycle marine battery and secure.  A fully charged
     battery should provide power for at least 2 sampling runs.  Depending on
     the battery available, more consecutive runs may be possible. Store the
     battery in a plastic container near the sampler to conceal and protect it
     from the weather.

     If collocated samplers will be located at the site, the two samplers must
     be within four (4) meters of each other, but outside of two (2) meters.
     The inlet heights must be within one (1) meter vertically.
OPERATING PROCEDURE

A.   Equipment and Supplies

     BGI PQ167R
     ERG Cr+6 filter holder apparatus
     Flow calibrator
     Logbook
     ERG filter assembly with glass funnel
     Cooler with ice substitute
     Powderfree gloves
     ERG sample paperwork

B.   Sampler and Sample Media Receipt Activities

     1.    Plug sampler into AC power and charge the internal battery for at
          least 24 hours.

     2.    Check parts and components against the packing list.

     3.    After charging, ensure sampler will power up and that the main
          screen is operational.
                           Page 9 of 19

-------
                               Toxics in Schools, Hexavalent Chromium SOP
                                                         July 30, 2009
     4.    The sampler may arrive with a default flow rate of 16.7 Lpm. If
          during the initial verification, the sampler's target flow rate is
          displayed as 16.7 Lpm, it must be changed to 15 Lpm.  Proceed
          directly to the calibration section of the SOP for direction  in making
          the change.

     5.    The ERG Cr+6 Filter Holder Modules will arrive to the field office  in
          a cooler with frozen ice substitutes.  The modules will have
          paperwork designating them for a specific site and run day. The
          modules must be kept in a freezer prior to sampling.  During
          transport to the monitoring site for run preparation, the filters must
          be kept cold as well. Samples must be returned to ERG cold
          using ice substitutes.

C.   Verification

     NOTE: THE PQ100 DOES NOT REQUIRE A LEAK TEST. CUTTING
     OFF THE FLOW OF AIR BY COVERING OR RESTRICTING THE AIR
     FLOW TO THE INLET WILL CAUSE DAMAGE TO THE INTERNAL
     PUMP AND WILL VOID THE WARRANTY.

     To VERIFY flow:

     1.    Install a test ERG Cr+6 filter holder module if available. If a test
          module is not available, the module to be used for the next sample
          day is acceptable; however, the module must be used immediately
          following the verification/calibration.

     2.    Attach a  NIST traceable flow standard to the inlet of the filter
          module.  Ensure the flow standard is on and has equilibrated to
          ambient conditions.

     3.    Turn on the PQ167R by pushing the "ON/OFF" button. If a
          message is blinking on the display, press "ENTER" to proceed to
          the "MAIN IDLE DISPLAY".

          The screen display should read:

          ETOOOOMin TSOO.OOM (Date)
          Q(Flow)Lpm T(Time) Bty(Capacity)%

          (Date) -today's date in military notation; e.g., 01JAN= January 1st
          (Flow) - the current flow rate selected to be regulated.
          (Time)-military time; e.g., 13:08= 13 Hours 8 Minutes or 1:08 PM
          (Capacity) - remaining charge in the internal battery.
                          Page 10 of 19

-------
                              Toxics in Schools, Hexavalent Chromium SOP
                                                        July 30, 2009
     4.   Press SETUP three times until the Set START DATE and TIME
         screen appears: The screen should appear as below:

         Set START DATE and TIME
         (Date)     (Time)     Off

     5.   The word, "Off, should be displayed in the lower right corner of the
         screen.  The bottom line of the display should be flashing.  If "On" is
         displayed, press the "ENTER" button until "On" stops flashing.
         Then toggle to "Off" by pressing the + or- buttons.

     6.   Press the "SETUP" button twice to get to the "MAIN IDLE
         DISPLAY"

     7.   Press the "RUN/STOP" button to activate the pump.

     8.   Allow  the pump to stabilize for at least 2 minutes.

     9.   If the measured flow and the flow indicated on the flow standard are
         within 4%, the sampler's calibration is acceptable.  If the flow is
         outside 4%, the unit must be recalibrated.

     10.  Press the "RUN/STOP" button to turn off the pump.

D.   Calibration

     NOTE: THE PQ100 DOES NOT REQUIRE A LEAK TEST. CUTTING
     OFF THE FLOW OF AIR BY COVERING OR RESTRICTING THE AIR
     FLOW TO THE INLET WILL CAUSE DAMAGE TO THE INTERNAL
     PUMP AND WILL VOID THE WARRANTY.

     To CALIBRATE flow:

     1.   Install a test ERG Cr+6 filter holder module if available.  If a test
         module is not available, the module to be used for the next sample
         day is acceptable; however, the module must be used immediately
         following the verification/calibration.

     2.   Press "SETUP". The screen will read; "Select FLOW RATE"

     3.   From  the "MAIN IDLE DISPLAY" press the "Setup" key once until
         the message below appears;

         Select FLOW RATE

         The Target Q should read 15.0 Lpm.  If it does not read 15.0 Lpm,
         set TARGET FLOW RATE to 15.0 Lpm by pressing ENTER.

                          Page 11 of 19

-------
                                Toxics in Schools, Hexavalent Chromium SOP
                                                          July 30, 2009
          The whole number value will remain on constant while the tenths
          still blink); use "+" or"-" to increase or decrease until 15 is
          displayed.  Press ENTER (Tenths value will now remain constant
          while whole number blinks); use "+" or"-" to increase or decrease
          until .0 is displayed.

     4.    From the "Select FLOW RATE" message screen, press both the
          "Reset" key and the "Run/Stop" key simultaneously to enter the
          calibration mode and the message below will appear;

          CALIBRATE Target=15.0 Lpm

     5.    Press the "RUN/STOP" button to activate the pump and the
          message below will appear:

          CALIBRATE Target = 15.0 Lpm
          Reference Q..        XX.X

          The Reference Q is an approximate flow rate used only as a visual
          aid in finding the corrected flow on the calibration device.  This value
          may indicate 5 to 15% error. This is for reference only!

     6.    Use the "+/-" keys to move the pump speed up or down until the
          calibration device indicates the desired flow rate.

     7.    When a stable reading has been  achieved, press the "ENTER" key
          to store the flow rate.

     8.    Exit the Setup menu and return to the "MAIN IDLE DISPLAY".
          CALIBRATIONS ARE NOT AFFECTED UNTIL THE ENTER KEY
          IS PRESSED AND THE PUMP IS RUNNING.

     9.    Record pre- and post- flow measurements and adjustments in the
          logbook.

E.   Conducting the Sampling Event

     Site Arrival Daily Activities

     1.    Visually inspect and ensure all 0-rings are in place and secure.
          Replace if necessary.

     2.    Always ensure that samples and  unused ERG Cr+6 Filter Holder
          Modules are transported  to and from the site cold.

     3.    Confirm all  cables (electrical connections) are secure, and that
          exterior connections are protected from the elements.

                          Page 12 of 19

-------
                           Toxics in Schools, Hexavalent Chromium SOP
                                                     July 30, 2009

4.    Record activities, site observations, and maintenance activities in
     logbook.

Preparing Sampler for a Sampling Event

1.    Prepare sample paperwork.  On the ERG AMBIENT HEXAVALENT
     CHROMIUM DATA SHEET, complete the "Lab Pre-Samp." and
     "Field Setup" sections. Record any pertinent observations in the
     notes section at the bottom of the form.

2.    Turn on the PQ167R by pushing  the "ON/OFF" button.  If a
     message is blinking on the display, press "ENTER" to proceed to
     the "MAIN  IDLE DISPLAY".  Then press "RESET" to clear prior run
     data.

3.    Conduct an initial flow check (verification) by following the
     instructions in section C. Verification.  Record the measurement
     from the flow standard on the ERG AMBIENT HEXAVALENT
     CHROMIUM DATA SHEET under the "Field Setup" section on the
     "Initial Rotameter Setting".

4.    Following the flow check, the screen display should read:

     ETOOOOMin TSOO.OOM (Date)
     Q(Flow)Lpm T(Time) Bty(Capacity)%

     (Date) -today's date in military notation; e.g., 01JAN= January 1st
     (Flow) - the current flow rate selected to be regulated.
     (Time)-military time; e.g., 13:08= 13 Hours 8 Minutes or 1:08 PM
     (Capacity)  - remaining charge in the internal battery.

5.    Press "SETUP".  The screen will  read; "Select FLOW RATE"
     The flow rate value will be blinking.

6.    The flow rate should read 15.0 Lpm. If it does not read 15.0
     Lpm, the unit must be calibrated to 15.0 Lpm. See calibration
     section for  adjusting target flow rate and calibration.

7.    Press "SETUP".  This is the  date and time screen.

     The screen should read;

     Set  DATE and TIME
     (dd) (mmm) (yyyy) (time)
                     Page 13 of 19

-------
                          Toxics in Schools, Hexavalent Chromium SOP
                                                    July 30, 2009
     To change the Date and Time;

     TIP: Only the field not blinking can be adjusted. Push enter to
         move to the next field.

     a.  DAY: Press ENTER and change by pressing the + or - key.
        When the day is correct, press ENTER.

     b.  MONTH: To change, press + or - key. When correct, press
        ENTER.

     c.  YEAR: To change, press + or - key. When correct, press
        ENTER.

     d.  TIME (hrs): To change, press + or - key. When correct, press
        ENTER.

     e.  TIME (min): To change,  press + or - key. When correct, press
        ENTER.

8.    When date and time are correct press "SETUP"

9.    This is the sample start screen which reads;

     Set START DATE and TIME
     (dd)(mmm) 00:00     Off

     This screen allows you to set a start date and time for a sampling
     run. The default is set to midnight the next day. To designate your
     own start date and time:

     a.  DAY: Press ENTER and change by pressing the + or - key.
        When the day is correct, press ENTER.

     b.  MONTH: To change, press + or - key. When correct, press
        ENTER.

     c.  YEAR: To change, press + or - key. When correct, press
        ENTER.

     d.  TIME (hrs): To change, press + or - key. When correct, press
        ENTER.

     e.  TIME (min): To change,  press + or - key. When correct, press
        ENTER.
                     Page 14 of 19

-------
                          Toxics in Schools, Hexavalent Chromium SOP
                                                   July 30, 2009
     f.   Enable the run by setting the "On/Off" function on the screen to
        "On".

        WARNING: The sampler will not automatically activate if
        this option is set to "Off".

10.   Press "SETUP"

     The screen will read;

     Set RUN TIME
     Hours: 24 Min: 00 On

     Set to 24 hours 0 minutes.  The default is always 24 hrs 0 min, the
     required sample duration. If the sample time needs to be modified,
     adjust as instructed in step 6 and 8.

11.   Press "SETUP".  The screen will return to the "MAIN IDLE
     DISPLAY"

     WARNING: DO NOT  PRESS THE  RESET BUTTON AT THIS
     TIME AS THE START TIME AND RUN TIME WILL DEFAULT.

12.   Press "RUN/STOP"

     If the START TIME ENABLE is set to "On" then the message
     "Alarm Triggered Run..."followed by "PQ100 Powering Down.."will
     appear briefly.  The PQ100 is now waiting for the internal real time
     clock to achieve the designated start time and will then power itself
     on and begin the  sampling run. If the START TIME ENABLE is set
     to "Off" then the pump will begin to run immediately. If this occurs,
     press RUN/STOP and begin back at step 2 ensuring START TIME
     ENABLE is set to "On".

Installing the ERG Cr+6 Filter Holder Module

     NOTE:  Gloves must be changed  for each sample, i.e. between
     retrieving a sample and preparing a new run gloves MUST be
     changed to prevent cross contamination.

1.    Remove the sample inlet cover on the stainless steel probe and
     make sure there is no  contamination on the probe.

2.    Put on a clean pair of powderfree gloves
                     Page 15 of 19

-------
                           Toxics in Schools, Hexavalent Chromium SOP
                                                    July 30, 2009
3.    Take the ERG Cr+6 Filter Holder Module storage container from the
     cooler and carefully remove the module.  The module may be in a
     plastic bag.  Return the bag to the container for use in the collection
     procedure.

4.    Make sure the glass funnel is securely attached to the filter holder.
     Loosen the small top nut on the filter container. Arrows will be
     present on the filter holder showing air flow direction and they
     should always point to the end of the sample probe line.

5.    Holding the module with the glass funnel facing down, slide the
     probe into the top fitting of the filter module and tighten the nut.
     Tighten the nut until the ERG Cr+6 Filter Holder Module is securely
     fastened to the probe.  Do not overtighten the plastic  nut.

     NOTE: If running  a field blank, repeat steps 1 through 5, count
     to 10, and then remove the field blank filter holder module and
     place it back into the antistatic bag.  Label the bag to designate
     the filter module as a field blank.  Log the filter ID as field blank
     in the comments section of the ERG Hexavalent Chromium
     Sample Data Sheet. The field blank must be run before the
     sample filter module is fastened to the probe.

Sample Recovery and Data Collection

     NOTE: Gloves must be changed for each sample, i.e. between
     retrieving a sample and preparing a new run, gloves MUST be
     changed to prevent cross contamination.

1.    On the ERG AMBIENT HEXAVALENT  CHROMIUM DATA SHEET,
     fill in  the "Field Recovery" section. Be sure to fill in the "Recovery
     Date", "Recovery Time", "Elapsed Time" (ETXXXXMin from
     sampler), and circle a "Status" selection.  This information will be on
     the "MAIN STATUS SCREEN".

2.    Conduct a final flow check (verification) by following the
     instructions in section C. Verification.  Record the measurement
     from  the flow standard on the ERG AMBIENT HEXAVALENT
     CHROMIUM  DATA SHEET under the "Field Setup" section, "Final
     Rotameter Reading".

3.    Put on a clean pair of powderfree gloves

4.    Take the module storage container from the cooler, open, and set
     aside.
                     Page 16 of 19

-------
                           Toxics in Schools, Hexavalent Chromium SOP
                                                      July 30, 2009
5.    While holding the ERG Cr+6 Filter Holder Module, loosen the top
     nut holding the module to the sample inlet and slide the module off
     the stainless steel probe.

6.    Place the ERG Cr+6 Filter Holder Module including glass funnel in
     the plastic bag and place back into the storage container. Place the
     storage container into a cooler with ice substitutes.

7.    Place cover back on end of probe line.

8.    Data may be downloaded to a laptop using the PQ100/200
     DOWNLOAD SOFTWARE.  ERG does not require this data, but
     direction can be found in the BGI PQ167 Quick Start document,
     pages 16 and 17.

Sample Shipping

The ERG Cr+6 Filter Holder Module container must be packed in a cooler
with ice substitutes and shipped overnight cold to ERG. The sample
paperwork must be included in the shipment. Use the pre-filled out
FedEx label provided by ERG, and fill out the "Sender" section with the
sampling agency's address and phone number.  Send priority overnight
to ERG.

If the shipping form is lost, use the address below for shipping to ERG,
and contact them directly for the FedEx accounting.

Address: ERG
         601 Keystone Park Drive
         Suite 700
         Morrisville, NC 27560
         919-468-7924
                      Page 17 of 19

-------
                                     Toxics in Schools, Hexavalent Chromium SOP
                                                               July 30, 2009
IV.   QUALITY ASSURANCE

     To ensure that quality data is being collected the following checks should be
     considered:

     A.   Flow Calibration

          A flow verification must be completed at the beginning of the study
          period. If the verification does not compare within 4%, the flow must be
          calibrated.  Document all quality assurance activities in the logbook.

     B.   Flow Verifications

          The flow must be verified or checked at the beginning and end of the
          sampling event to determine an average sample flow. Document all
          quality assurance activities and observations in the logbook.

     C.   Independent Audits

          If possible, it is recommended that an independent flow check of the
          sampler be conducted at some point during the study. This check may
          be conducted by a state or local agency's quality assurance team or
          independent audit program.

V.   DATA FORMS

     All sample  related run data forms will be supplied by ERG. Check the data
     sheets for completion after every setup or retrieval event.  The operator is
     expected to keep a logbook to document all site activities, quality assurance
     activities, and sampling activities. The ERG AMBIENT HEXAVALENT
     CHROMIUM DATA SHEET is attached below.
                               Page 18 of 19

-------
                                            Toxics in Schools, Hexavalent Chromium SOP
                                                                             July 30, 2009
              ERG Ambient Hexavalent Chromium Sample Data Sheet
                                                    ERG Lab ID #
                    AMBIENT HEXAVALENT CHROMIUM DATA SHEET
     O)
     £
    Q.
        Site Code:
        City/State:
        AQS Code:
                                       Collection Date:
                                       Primary Event (Y/N):
                                       Collocated Event (Y/N):
   Q,
   I
   OT
Site Operator:
Set-Up Date:
Collection Date:
Batch  I.D. No.:
                                                       System #:
                  Elapsed Timer Reset (Y/N):
        Initial Rotameter Setting (C.O. B.):
        Programmed Start Time:  	
                                                       (After 5 minutes warm-up)
                                           Programmed End Time:
     t
  •o  «
  "5  o
  11  g
    cc
Recovery Date:  	
Final Rotameter Reading (C.O.B.):
Elapsed Time:  	
                 Recovery Time:
                  Status:
_(After 5 minutes warm-up)
 Valid      Void   (Circle one)
   I
   O
   o
   
-------