United States        Office of Research and     EPA/600-R-01/088
          Environmental Protection    Development        September 2001
          Agency          Washington DC 20460

«€PA     Innovative Technology
          Verification Report

          Field  Measurement
          Technologies for Total
          Petroleum Hydrocarbons in Soil

          Wilks Enterprise, Inc. Infracal ®
          TOG/TPH Analyzer

-------
                    UNITED STATES ENVIRONMENTAL PROTECTION AGENCY
                                         Office of Research and Development
                                              Washington, DC 20460
                    Environmental Technology Verification Program

                                       Verification Statement


TECHNOLOGY TYPE: FIELD MEASUREMENT DEVICE

APPLICATION: MEASUREMENT OF TOTAL PETROLEUM HYDROCARBONS

TECHNOLOGY NAME: Infracal® TOG/TPH ANALYZER

COMPANY: WILKS ENTERPRISE, INC.
ADDRESS: 140 WATER STREET
            SOUTH NORWALK, CT 06854

WEB SITE: http://www.wilksir.com

TELEPHONE: (203) 855-9136

VERIFICATION PROGRAM DESCRIPTION

The U.S. Environmental Protection Agency (EPA) created the Superfund Innovative Technology Evaluation (SITE) and
Environmental Technology Verification (ETV) Programs to facilitate deployment of innovative technologies through
performance verification and information dissemination. The goal of these programs is to further environmental protection
by substantially accelerating the acceptance and use of improved and cost-effective technologies. These programs assist and
inform those involved in design, distribution, permitting, and purchase of environmental technologies. This document
summarizes results of a demonstration of the Infracal® TOG/TPH Analyzer developed by Wilks Enterprise, Inc. (Wilks).

PROGRAM OPERATION

Under the SITE and  ETV Programs, with the full participation of the technology developers, the EPA evaluates and
documents the performance of innovative technologies by developing demonstration plans, conducting field tests, collecting
and analyzing demonstration data, and preparing reports.  The technologies are evaluated under rigorous quality assurance
(QA) protocols to produce well-documented data of known quality. The EPA National Exposure Research Laboratory, which
demonstrates field sampling, monitoring, and measurement technologies, selected Terra  Tech EM Inc. as the verification
organization to assist in field testing seven field measurement devices for total petroleum hydrocarbons (TPH) in soil. This
demonstration was funded by the SITE Program.

DEMONSTRATION DESCRIPTION

In June 2000, the EPA conducted a field demonstration of the Infracal® TOG/TPH Analyzer and six other field measurement
devices for TPH  in soil. This verification  statement focuses on the Infracal® TOG/TPH Analyzer; a similar statement has
been prepared for each of the other six devices. The performance and cost of the Infracal® TOG/TPH Analyzer were
compared to those of an off-site laboratory reference method, "Test Methods for Evaluating Solid Waste" (SW-846)
Method 8015B (modified). To verify a wide range of performance attributes, the demonstration had both primary and
secondary objectives. The primary objectives included (1) determining the method detection limit, (2) evaluating the
accuracy and precision of TPH measurement, (3) evaluating the effect of interferents, and (4) evaluating the effect of moisture
content on TPH measurement for each device. Additional primary objectives were to measure sample throughput and
estimate TPH measurement costs. Secondary objectives included (1) documenting the skills and training required to properly
operate the device, (2) documenting the portability of the device, (3) evaluating the device's durability, and (4) documenting
the availability of the device and associated spare parts.

-------
The InfracalisTOG/TPH Analyzer was demonstrated by using it to analyze 74 soil environmental samples, 91 soil
performance evaluation (PE) samples, and 50 liquid PE samples. The environmental samples were collected in five areas
contaminated with gasoline, diesel, lubricating oil, or other petroleum products, and the PE samples were obtained from a
commercial provider. Collectively, the environmental and PE samples provided the different matrix types and the different
levels and types of petroleum hydrocarbon (PHC) contamination needed to perform a comprehensive evaluation of the
Infracak>TOG/TPH Analyzer. During the demonstration, Wilks analyzed most of the  samples using the device equipped with
one of two  sample stages: Model CVH or Model HATR-T. Only 8 percent of the samples were analyzed using both models.

In general,  Model CVH was used to analyze samples containing gasoline range organics (GRO) and Model HATR-T was
used to analyze samples that did not contain GRO. For this reason, the performance of the Infracal®TOG/TPH Analyzer as
a whole was evaluated, but that of each model was not. A complete description of the demonstration and a summary of its
results are available in the "Innovative Technology Verification Report: Field Measurement Devices for Total Petroleum
Hydrocarbons in Soil—Wilks Enterprise, Inc., InfracaLsTOG/TPH Analyzer" (EPA/600/R-01/088).

TECHNOLOGY DESCRIPTION

The Infracal®TOG/TPH Analyzer developed by Wilks is based on infrared analysis. The device can be operated  as either
Model CVH or Model HATR-T simply by switching sample stages. Model CVH uses a sample stage that contains a quartz
cuvette, and Model HATR-T uses the cubic zirconia horizontal attenuated total reflection sample stage. Model CVH is used
when a sample contains GRO, extended diesel range organics (EDRO), or both, and Model HATR-T is used when a sample
contains only EDRO. Because of the environmental hazards associated with chlorofluorocarbons, Model HATR-T, which
uses Vertrel®MCA, is preferred over Model CVH, which uses  Freon 113, a chlorofluorocarbon. However, according to
Wilks, Model CVH is more sensitive and can achieve a lower detection limit than Model HATR-T.

The Infracal®TOG/TPH Analyzer includes a single-beam, fixed-wavelength, nondispersive infrared filter-based
spectrophotometer with a dual detector system. In Model CVH, a pulsed beam of infrared radiation from a tungsten lamp
is transmitted through a quartz cuvette that contains a sample extract. In Model HATR-T, which is based on an evaporation
technique, an extract is placed directly on the sample stage. The radiation that passes through the extract enters the dual
detector system, whose filters isolate a reference wavelength (2,500 nanometers) and an analytical wavelength
(3,400 nanometers) to measure PHCs present in the extract.

During the  demonstration, Wilks first dried a given soil sample by adding silica gel. Extraction of PHCs from the sample
was typically performed by adding 20 milliliters of Freon 113 (for Model CVH) or Vertrel®MCA (for Model HATR-T) to
20 grams of the sample. The mixture was agitated by means of vigorous shaking, and  the sample extract was decanted into
an extraction reservoir. Using an air syringe, Wilks filtered the extract (1) into a quartz cuvette that was placed in
Model CVH or (2) into a beaker and then transferred the extract to the center of the HATR-T sample stage using  a
micro syringe. Finally, Wilks read the TPH concentration in milligrams per kilogram on a digital display.

VERIFICATION OF PERFORMANCE

To ensure data usability, data quality indicators for accuracy, precision, representativeness, completeness, and comparability
were assessed for the reference method based on project-specific QA objectives. Although the reference method results
generally exhibited a negative bias, based on the results for the data quality indicators, the reference method results were
considered to be of adequate quality. The bias was considered to be significant primarily for low-  and mediumconcentration-
range soil samples containing diesel, which made up only 13 percent of the total number of samples analyzed
during the demonstration. The reference  method recoveries observed during the demonstration were typical of the recoveries
obtained by most organic analytical methods for environmental samples, hi general, the user should exercise caution when
evaluating the accuracy of a field measurement device by comparing it to reference methods because the reference methods
themselves may have limitations. Key demonstration findings are summarized below for the primary and secondary
objectives.

Method Detection Limit: Based on the TPH results for seven low-concentration-range diesel soil  PE samples, the method
detection limits were determined to be 76 and 4.79 milligrams  per kilogram for the Infracak>TOG/TPH Analyzer
(Model HATR-T) and reference method, respectively.

Accuracy and Precision: Seventy-two of 101 Infracal®TOG/TPH Analyzer results (71 percent) used to draw conclusions
regarding whether the TPH concentration in a given sampling area or sample type exceeded a specified action level agreed
with those of the reference method; 2 device conclusions were  false positives, and 27 were false negatives.

-------
Of 105 InfracalisTOG/TPH Analyzer results used to assess measurement bias, 22 were within 30 percent, 28 were within
30 to 50 percent, and 55 were not within 50 percent of the reference method results; 78 device results were biased low, and
27 were biased high.

For soil environmental samples, the Infracal®TOG/TPH Analyzer results were statistically (1) the same as the reference
method results for one of the five sampling areas and (2) different from the reference method results for four sampling areas.
For soil PE samples, the device results were statistically different from the reference method results for medium- and
highconcentration-range weathered gasoline and diesel samples. For liquid PE samples, the device results were statistically
different from the reference method results for both weathered gasoline and diesel samples.

The InfracaleTOG/TPH Analyzer results correlated highly with the reference method results for two of the five sampling
areas and weathered gasoline soil PE  samples (the square of the correlation  coefficient [R2] values ranged  from 0.85 to 0.94,
and F-test probability values were less than 5 percent). The device results correlated moderately with the reference method
results for two sampling areas and diesel soil PE samples (R2values ranged  from 0.59 to 0.68, and F-test probability values
were less than 5 percent). The device results correlated weakly with the reference method results for one sampling area (the
R2 value was 0.14, and the F-test probability value was 35.32 percent).

Comparison of the Infracak>TOG/TPH Analyzer and reference method median relative standard deviations (RSD) showed
that the device exhibited less overall precision than the reference method. Specifically, the median RSD ranges were 5 to
30 percent and 5.5 to 18 percent for the device and reference method, respectively.

Effect of Interfere nts:  The Infracak>TOG/TPH Analyzer showed a mean response of less than 1 percent for neat
tetrachloroethene (PCE); neat 1,2,4-trichlorobenzene; and soil spiked with humic acid. The device's mean responses for neat
methyl-tert-butyl ether (MTBE), Stoddard solvent, and turpentine were 62,  120, and 77 percent, respectively. The reference
method showed varying mean responses for MTBE (39 percent); PCE (17.5 percent); Stoddard solvent (85 percent);
turpentine (52 percent); 1,2,4-trichlorobenzene (50  percent); and humic acid (0 percent). For the demonstration, MTBE and
Stoddard solvent were included in the definition of TPH.

Effect of Moisture Content: Soil moisture content had a statistically significant impact on the Infracal®TOG/TPH Analyzer
TPH results for diesel soil PE samples but not on those for weathered gasoline soil PE samples. Specifically, the device
showed a three-fold increase in TPH results for diesel samples when the soil moisture content was increased from less than
1 percent to 9 percent. The reference  method TPH results were unaffected when the soil moisture content was increased.

Measurement Time: From the time of sample receipt, Wilks required 35 hours, 30 minutes, to prepare a draft data package
containing TPH results for 215 samples compared to 30 days for the reference method.

Measurement Costs: For the Infracal®TOG/TPH Analyzer, the TPH measurement cost for 21 5 samples was estimated to
be $6,450  (including the monthly rental cost of the device, whose purchase price is $6,200) compared to $44,410 for the
reference method.

Skill and Training Requirements: The Infracak>TOG/TPH Analyzer can be operated by one person with  basic wet chemistry
skills. The sample analysis procedure for the device can be learned in the field with a few practice attempts. During the
demonstration, some of the items used during the sample preparation procedure made the TPH measurement procedure less
simple and more time-consuming.

Portability: The Infracak>TOG/TPH Analyzer can be easily moved between sampling areas in the field, if necessary. It can
be operated using a  110-volt alternating current power source or a direct current power source.

Durability and Availability of the Device: During a 1 -year warranty period, if the infrared spectrophotometer or a sample
stage malfunctions,  Wilks  will provide a replacement item within 48 hours on loan for a fee of $75 while the original item
is being repaired. During the demonstration, Model CVH proved to be durable and did not malfunction or become damaged.
However,  the spectrophotometer malfunctioned when the Model CVH sample stage was replaced with the Model HATR-T
sample stage. Wilks does not supply some items necessary for TPH measurement using the device (for example, extraction
solvents). The availability of replacement or spare parts not supplied by Wilks depends on their manufacturer or distributor.

In summary, during  the demonstration, the Infracak>TOG/TPH Analyzer exhibited the following desirable characteristics
of a field TPH measurement device: (1) sensitivity to interferents that are PHCs (MTBE and Stoddard solvent), (2) lack  of
sensitivity to interferents that are not PHCs (PCE; 1,2,4-trichlorobenzene; and humic acid), (3) high sample throughput,  and
(4) low measurement costs. However, the device TPH results did not compare well with the reference method results. In

-------
addition, turpentine biased the device TPH results high, indicating that the accuracy of TPH measurement using the device
will likely be impacted by naturally occurring oil and grease present in soil that are not removed by silica gel. Also, the
device TPH results for diesel soil PE samples showed a three-fold increase when the soil moisture content was increased by
8 percentage points.  Finally, the device results obtained using the two  sample stages did not agree. Collectively, these
demonstration findings indicated that the Infracak>TOG/TPH Analyzer may be considered for TPH screening purposes;
however, the user should exercise caution when considering the device for a field TPH measurement application requiring
definitive results.

Original
signed by

Gary J. Foley, Ph.D.
Director
National Exposure Research Laboratory
Office of Research and Development

NOTICE: EPA verifications are based on an evaluation of technology performance under specific, predetermined criteria and
appropriate quality assurance procedures. The EPA makes no expressed or implied warranties as to the performance of the technology
and does not certify that a technology will always operate as verified. The end user is solely responsible for complying with any and
all applicable federal, state, and local requirements.

-------