School Air Toxics, SVOC / PAH SOP
                                         July 30, 2009
     STANDARD OPERATING PROCEDURE
                     FOR
 MEASUREMENT OF SEMI-VOLATILE ORGANIC
           COMPOUNDS USING THE
TISCH ENVIRONMENTAL INC. TE-PUF SAMPLER
           U.S. Environmental Protection Agency
       Region 4, Science and Ecosystem Support Division
               Athens, Georgia, 30605
                  Page 1 of 15

-------
                                               School Air Toxics, SVOC / PAH SOP
                                                               July 30, 2009
                            Acknowledgement

This Standard Operating Procedure (SOP) was developed by EPA Region 4,
Science and Ecosystem Support Division.  This SOP is based in part on the
Commonwealth of Kentucky's ambient monitoring SOP, Measurement of Semi-
Volatile Organic Compounds by the PDF Sampler (TS17-16). In addition, Eastern
Research Group (ERG), Morrisville, North Carolina contributed greatly in the
development of this document. Special thanks to the Commonwealth of Kentucky,
Tisch Environmental, Inc. and ERG for sharing their work in developing this SOP for
the National Toxics in Schools Initiative.

For questions or comments please contact:

Tim Slagle, EPA, Region 4, SESD at slagle.tim(S)epa.gov or 706-355-8741
Greg Noah, EPA, Region 4, SESD at noah.greg@epa.gov or 706-355-8635
Mike Jones, EPA-OAQPS-AQAD at iones.mike(S)epa.gov or 919-541-0528
                              Page 2 of 15

-------
                                                    School Air Toxics, SVOC / PAH SOP
                                                                     July 30, 2009
                               Table of Contents

I.  Introduction	4
II.  Installation	7
      A. Sampler Installation and Siting	7
III. Calibration	7
      A. Equipment and Supplies	7
      B. Procedure	7
IV.    Standard Operating Prodedure	8
      A. Equipment and Supplies	8
      B. Sample Delivery to Field	8
      C. Collection Event Set-up and Loading the PS-1 Sampling Head	9
      D. Collection Event Set-up and Setting up the PS-1	10
      E. Collection Event Recovery	11
      F. Collection Event Recovery and Unloading the Sampling Head	11
V.  Preventative Maintenance	14
      A. Schedule	14
VI. Data Forms	14
      A. ERG PM10 Metals Sample Data Sheet	15
                                  Page 3 of 15

-------
                                              School Air Toxics, SVOC / PAH SOP
                                                               July 30, 2009
INTRODUCTION

This method is designed to simultaneously collect suspended airborne
particulates as well as trap Semi-Volatile Organic Compounds (SVOCs) such as
Polycyclic Aromatic Hydrocarbons (PAHs).

The Poly-Urethane Foam sampler (Figure 1), commonly known as the "PUF
sampler" is composed of an aluminum shelter, dual chamber sampling head, flow
venturi, magnehelic gauge, voltage variator / elapsed time indicator, blower
motor, exhaust hose, and seven-day skip timer.

The dual chamber sampling head (Figure 2), contains both filtering systems. The
upper chamber supports the 4" diameter Micro-Quartz particulate filter while the
lower chamber houses the glass cartridge that contains the PolyUrethane Foam
and XAD-2® for vapor entrapment.

The blower motor is manufactured with a by-pass and cooling fan to permit the
motor to operate at low sampling flow rates for periods of long duration without
motor failure from overheating. Optimum flow rate for the sampler is 0.225
m3/min. which yields a total volume of air sampled per 24 hours of greater than
300 m3.

The voltage variator adjusting screw alters the blower motor speed to achieve the
desired flow rate. The air flow rate is measured through the flow venturi utilizing
a 0-100" magnehelic gauge.

A seven-day skip timer is used to automatically initiate and terminate sampling at
preset times.  The actual length of sampling is measured by the elapsed time
indicator. The whole assembly is housed in a specially designed aluminum
shelter that provides protection from precipitation while maintaining ample space
for air entering or leaving the sampler.  The height of the shelter puts the air inlet
at about one meter above  the grade.

This SOP is designed to be a  step by step method for operating the sampler to
be used in conjunction with the manufacturer's operations manual;
http://www.tisch-env.com/tisch/pdf/te1000.PDF.  Maintenance and
troubleshooting should be  conducted using the Tisch; TE-PUF operations
manual.

Laboratory Analysis Methodology may be referenced by contacting the Eastern
Research Group (ERG) directly at 919-468-7800 or by email;
Julie.Swift(S)erg.com.
                            Page 4 of 15

-------
                                                 School Air Toxics, SVOC / PAH SOP
                                                                  July 30, 2009
           FIGURE 1. Polyurethane Foam (PUF) Sampler
Magnenlic Gauge
   0-100 in.
   Exhaust Duct
   (6 in. x 10 ft)
                       Sampling Head
                        see Figure 3)
                                                                Voltage Variator
                                                                 Elapsed Time
                                                                    Meter
                                                                  7-Day Timer
                              Page 5 of 15

-------
                                                 School Air Toxics, SVOC / PAH SOP
                                                                    July 30, 2009
                  FIGURE 2. Sampling Head
- LOWER CANISTER
                      GLASS CARTRIDGE AND
                     -ADSORBENT
              StLICONE RUBBER '
                GASKET
                                                       FILTER HOLDER WITH
                                                        SUPPORT SCREEN
                                                         4" DIAMETER FILTER
                                         FILTER RETAINING RING  /
                            Page 6 of 15

-------
                                                     School Air Toxics, SVOC / PAH SOP
                                                                      July 30, 2009
II.   INSTALLATION

     A.   Sampler Installation and Siting

          Assemble the sampler as described in the Tisch Environmental, Inc. TE-PUF
          Operations Manual shipped with the sampler which can also be found at;
          http://www.tisch-env.com/tisch/pdf/te1000.PDF.

          Check the sample site areas for safety. Ensure there will be enough room
          for the operator to move freely while working, and ensure physical conditions
          of the location will allow the operator to work safely.

          The sampler should be set in a location unobstructed from any side. No tree
          limbs or other hanging obstructions should be above the sampler.  It is
          suggested that the horizontal distance from the sampler to the closest
          vertical obstruction higher than the sampler be at least twice the height of
          the vertical obstruction. The exhaust hose should be stretched out in a
          down wind direction if possible.

          Install the sampler on a reasonably level structure at a height between two
          (2) and fifteen (15) meters above the ground.

          Two skids of approximately one meter length may be attached parallel to the
          legs of the sampler, and concrete blocks or sand bags may be placed on the
          skids to prevent tipping of the sampler.

          Connect the sampler to a grounded electrical outlet with 115 volts, and at
          least 15 amp service.  Protect the connector from precipitation by placing it
          in the shelter or wrapping it with plastic tape.

          If collocated samplers will be located at the site,  the two samplers must be
          within four (4) meters of each other,  but outside of two (2) meters.  The inlet
          heights must be within one (1) meter vertically.
III.  CALIBRATION

A.   Equipment and Supplies

          Empty glass cartridge
          Calibration kit containing; Manometer and orifice calibration unit

B.   Procedure

          For accurate data the hardware used to obtain this data must be checked
          and calibrated. This includes the Sample Train which consists of the Dual

                                  Page 7 of 15

-------
                                                    School Air Toxics, SVOC / PAH SOP
                                                                    July 30, 2009
          Chamber Sampling Head Flow through venturi and also the magnehelic
          gage. Calibration of the sampler is performed without a foam plug or filter
          paper in the sampling head. However the empty glass cartridge must
          remain  in the head to insure a good seal through the head.

          The procedure for calibration is described in the Tisch TE-PUF Operations
          Manual shipped with the sampler which can also be found at;
          http://www.tisch-env.com/tisch/pdf/te1000.PDF. The sample train must be
          calibrated with an orifice calibration unit.  The orifice calibration unit itself has
          been calibrated against a positive displacement primary standard (Roots
          Meter) so that it can be used to calibrate the sample train. The calibration
          data sheet for the orifice calibration unit should be copied for safe keeping.
          Data from this sheet will be required for calibration of the TE-PUF sampler.

          Operators collecting SVOC samples should be familiar with the methodology
          and calculations contained in EPA Compendium Method TO-13A
          Determination of Polycyclic Aromatic Hydrocarbons  (PAHs) in Ambient
          Air Using Gas Chromatography/Mass Spectrometry (GC/MS), which can
          be found at; http://www.epa.gov/ttn/amtic/files/ambient/airtox/to-13arr.pdf.

          The orifice is generally calibrated once a year. However, before each use, it
          should be inspected for visible signs of damage.  A calibration check must
          be made anytime the unit appears to have any nicks or dents.
IV.   STANDARD OPERATING PROCEDURE

     A.   Equipment and Supplies

          Logbook
          Tubular Glass PUF/XAD-2 resin cartridge
          110 mm diameter filter
          Powderless gloves (example; VWR®MICROGRIP® PURPLE NITRILE®
          Poly-Coated Powder-Free Gloves, Catalog # 40101-XXX for~$30 pkg. 100)
          ERG sample paperwork

     B.   Sample Delivery to Field

          The filter is packaged in a plastic petri dish sealed by wrapping it with Teflon
          tape and placed in a zip lock bag along with the associated ASVOC Sample
          Data Sheets (i.e., Chain of Custody (COC)) form. The ends of the glass
          cartridge are sealed with Teflon/0-ring press-in caps, and the entire
          cartridge is then wrapped in cleaned aluminum foil. The wrapped cartridge
          is then inserted into a bubble wrap bag and packaged in a plastic screw lid
          jar.  The plastic jar and  the zip lock bag containing the petri dish and COC
          are placed in a small cooler with a Blue Ice brick(s) and shipped to the field

                                 Page 8  of 15

-------
                                                School Air Toxics, SVOC / PAH SOP
                                                                 July 30, 2009
     site.  Upon receipt at the site, the filter and the plastic jar should be removed
     from the cooler and placed in a refrigerator with a maximum operating
     temperature of 4BC,  or in a freezer, until ready for use.  The Blue Ice brick(s)
     should be placed in a freezer until ready for re-use.

C.   Collection Event Set-up and Loading the PS-1 Sampling Head

     SETUP MUST BE PERFORMED ON THE DAY PRIOR TO THE
     SAMPLING  EVENT

     The PS-1 Sampling Head can be loaded at the laboratory/office or at the
     field site on the discretion of each program participant (agency).  Regardless
     of which approach is used, great care must be exercised to minimize the
     potential for  contaminating  the collection media. Whether the media is
     loaded at the laboratory/office or at the field site, it must be maintained at a
     cool temperature while transporting the media to the site location for set-up.
     The following instructions apply to loading the collection media into the PS-1
     sampling head at an alternate location in the event of adverse weather
     conditions:

     1.    Put on a pair of powder-free nitrile gloves

     2.    Separate the components of the sampling head.  Unscrew the
          Cartridge Holder from the Filter Holder Base by rotating in a counter-
          clockwise direction.  Detach the Filter Retaining Ring from the Filter
          Holder by loosening the wing nuts on pivot screw posts and rotating
          them downward from their associated tabs.  Place the components on
          a clean surface.

     3.    Remove the prepared sampling glass cartridge from it=s transport
          packaging.  Place the bubble wrap bag  back in the plastic jar.  Remove
          the aluminum foil wrap from around the cartridge.  Neatly fold the foil
          and place it back in the plastic jar.  Remove the Teflon/0-ring press-in
          caps from the inlet and outlet ports of the glass cartridge and place
          them back in the plastic jar.

     4.    Visually ensure that a silicone gasket is present in the bottom of the
          Cartridge Holder.  Insert the glass cartridge into the Cartridge Holder
          ensuring  that the fritted end is pointing  downward.

     5.    Visually ensure that a silicone gasket is present in the outlet end of the
          Filter Holder.  Place the Filter Holder on top of the Cartridge Holder
          and screw the 2 components together by rotating the Filter Holder in a
          clockwise direction until the silicone gaskets  barely seat against the
          lips of the glass cartridge. At this point, rotate the Filter Holder just a
          small amount further to seal the components; DO NOT OVER-

                              Page9of 15

-------
                                                School Air Toxics, SVOC / PAH SOP
                                                                 July 30, 2009
          TIGHTEN (over-tightening can cause the lips of the glass cartridge to
          chip or the cartridge to crack).

     6.    Remove the Teflon tape from the outside of the petri dish. Using
          cleaned Teflon coated forceps, gently pick up the filter media and
          center it on the Filter Support Screen positioned at the inlet of the Filter
          Holder.  Immediately place the Teflon Gasket and Filter Retaining Ring
          on top of the loaded filter, taking care to align the pivot screw posts on
          the Filter Holder with the corresponding tabs on the Filter Retaining
          Ring.  Ensure the filter is installed between the upper and lower
          Teflon® gaskets. Windy conditions can make this step tricky; try to
          replace the hold down ring as quickly as possible so the filter is not
          lost. If the filter is blown off the sampler; replace with a new filter. Lift
          the pivot screw posts up into position and tighten the wing nuts until
          they barely seat against the tabs. At this point, tighten each wing nut a
          small  amount and then move to the next one. Continue this process
          until all of the wings nuts are hand tight.

     7.    If you are  assembling the Sampling Head at the Laboratory/office, you
          will need to install the filter protection cap as part of Step 6, and then
          remove it  at the site. You will also want to cover the outlet port of the
          Sampling  Head with cleaned aluminum foil.

     8.    Prepare sample paperwork. On the ERG SVOC Sample Data  Sheet,
          complete the "Field Setup" section which includes "Site Operator",
          "Setup Date", "Collection Date", "System #", and "Elapsed Timer
          Reset" fields.

D.   Collection  Event Set-up and Setting up the PS-1

     1.    Remove the pin from the latch on the top of the sampler housing and
          open the top sampler cover/inlet. The top can be latched in  the open
          position from the rear.

     2.    Insure the throat quick-connect fitting gasket is installed correctly and
          in good condition.

     3.    Place the  sampling head quick-connect fitting into the throat quick-
          connect fitting, using the levers, lock the connection.  Ensure that the
          levers are parallel with the sampling head and the "rings" do not
          prevent the levers from completely closing.

     7)    Press the reset button (if equipped) on the elapsed time meter on the
          sampler and record the reading of the timer on the ERG SVOC Sample
          Data Sheet, under the "Elapsed Time, Start" section.  The meter
          should reset to all zeroes. If the timer is not equipped with a reset

                             Page 10 of 15

-------
                                              School Air Toxics, SVOC / PAH SOP
                                                               July 30, 2009
          button record the current reading of the timer.

     8)    On the ERG SVOC Sample Data Sheet, record the ambient
          temperature (°C) and barometric pressure (inches Hg) under the
          "Elapsed Time, Start" section.

     9)    Check the zero reading on the magnehelic gauge. Close the roof of the
          shelter and run the sampler for about 5 minutes to warm up the motor.
          While the motor warms up, read and record the ambient temperature
          and pressure.

     10)   Observe and record the magnehelic gauge reading on the PUF
          Sample Form (Figure 3) and in the station logbook in the space
          marked "Magnehelic Reading: Initial:". Note: the initial reading
          should correspond to a flow rate of 0.226 m3/min ±10%. If reading
          is outside of these limits adjust the ball valve or voltage variator
          screw for proper flow rate.

     11)   Turn off the sampler and set the seven-day timer to sample from
          midnight to midnight local time of the sampling day.

     12)   Record the Sample I.D., Batch #, Sampler I.D., Site I.D., County,
          Operator, Sample Date, and running time (elapsed) meter reading on
          the PUF Sample Form and in the logbook. (Be sure to record running
          time (elapsed) meter reading in the "Time: (Minutes), Initial" space.)

E.   Collection Event Recovery

     1. After the sampling period,  record the running time  (elapsed) meter
       reading in  the "Time: (Minutes),  Final" space  on  the form and in  the
       logbook. Check magnehelic  gauge zero.  Also read and record  the
       ambient temperature and pressure.

     2. Turn the sampler on and allow the motor to warm up for 5 minutes, then
       observe  and record the  magnehelic gauge  reading on  the PUF  Sample
       Form  and  in  the station logbook  in the  space marked "Magnehelic
       Reading: Final:".

F.   Collection Event Recovery and Unloading the Sampling Head

     The PS-1 Sampling  Head can be unloaded at the laboratory/office or at the
     field site on the discretion of each program participant (agency). Regardless
     of which approach is used, great care must be exercised to minimize the
     potential for contaminating the collection media when handling it. If the
     sampling head is removed from the PS-1 and transported to the laboratory
                            Page 11 of 15

-------
                                           School Air Toxics, SVOC / PAH SOP
                                                            July 30, 2009
or office to replace the collection media, it must remain in an upright position.
This is to prevent particulate collected on the filter from falling off.

The following steps are associated with unloading the collection media from
the PS-1 sampling head:

1)   Put on a pair of powder-free nitrile gloves. Note: A new set of gloves
     should always be used between handing parts of the sampling head
     (wing-nuts, the retaining ring, etc.) and the sampling media.

2)   Unscrew the Cartridge Holder from the Filter Holder Base by
     rotating in a counter-clockwise direction and place it on a clean
     surface. Open the petri dish.  Detach the Filter Retaining Ring from
     the Filter Holder by loosening the wing nuts on pivot screw posts and
     rotating them downward from their associated tabs.  Lift the Filter
     Retaining  Ring and remove the filter sample from the Filter Holder
     using cleaned Teflon coated forceps and place it in the open petri dish.
     With a new set of clean gloves on, gently grasp the filter sample and
     fold it in half so that the collected particulate from both halves is
     pressed against each other.  Now fold the filter in half again (i.e., quad-
     fold) and place it in the petri dish.  Re-seal the petri dish by wrapping it
     with new Teflon tape.

3)   Remove the retained aluminum foil, Teflon/0-ring press-in caps, and
     bubble wrap bag from plastic transport jar.

4)   Remove the glass cartridge sample from the Cartridge Holder.  Insert
     the Teflon/0-ring press-in caps into the cartridge inlet and outlet ports.
     Wrap the glass cartridge sample in the retained aluminum foil, and
     place it in  the bubble wrap bag.  Slide the bubble wrap bag (with the
     cartridge sample) into the transport plastic jar and screw on the cap.

5)   Using the  current calibration for the motor and the appropriate
     temperature range, convert the magnehelic readings to flow rate and
     record on  the form and in the logbook.

     NOTE:  Final flow rate must be within ±10% of the initial flow rate,
     otherwise the sample is to be flagged.

6.    Average the initial and final flow rates and record the average on the
     form and in the logbook.
7.    Subtract the initial running time meter reading from  the final reading
     and record the hours sampled in the appropriate space on the form
     and in the logbook.

     NOTE:  Hours sampled must be between 23 to 25 hours.

                        Page 12 of 15

-------
                                           School Air Toxics, SVOC / PAH SOP
                                                            July 30, 2009
8.    If the sample is defective, void it at this time by writing with big letters
     VOID on the form along with a brief reason for voiding.  Also enter
     reason for voiding in the logbook, for example - power off, motor dead,
     timer malfunction.

9.    For data validation purposes the site operator must record information
     concerning weather conditions and  unusual  site events that occur on
     the sampling date.

     The following information should be recorded in the appropriate spaces
     on the record forms for transmittal with the samples.

           Weather Conditions (sample date only).

           Report newspaper, local airport report, and/or own observations
           over the 24-hour sampling period.   The report should include:
           Temperatures (high-low), Precipitation (rain, snow,  etc.),   Sky
           (clear/cloudy),  Wind  (calm/light/moderate/strong),   and  Wind
           Direction.

           Unusual Events

           Report  fires,  tornadoes, dust storms,  construction,  demolition,
           etc., that may have an  impact at the site on the sampling date.
           Use newspaper  reports  and own observations.    The report
           should  include;  type  of event, sites in the area of the event,
           distance and direction from site, date of occurrence, pictures of
           the event if visible from site, weekly observations if event is on-
           going, and date event ended.

10.  At least one field filter/PUF blank must be returned to the laboratory
     with each group of samples. A field blank is treated exactly as a
     sample  except that no air is drawn through the filter/PUF cartridge
     assembly.

11.  Place the completed COC and the sealed petri dish into a zip lock bag.
     Place the zip lock bag and the transport plastic jar into the cooler
     initially used to ship the media to the site. Place the frozen Blue Ice
     brick(s)  in the cooler and ship the cooler back to the ERG Laboratory
     using overnight service.  If the Sampling Head is transported back to
     the laboratory/office for recovery, the media must be maintained at a
     cool temperature while transporting, and protected from ultraviolet (UV)
     light to prevent possible photo-decomposition of collected analytes.
     Also,  care must be taken to insure that the Sampling Head is
                        Page 13 of 15

-------
                                                School Air Toxics, SVOC / PAH SOP
                                                               July 30, 2009
              transported in a vertical position with the filter pointing upward to
              ensure that particulate matter is not dislodged from the filter and lost.

ALL SAMPLES MUST BE SHIPPED AS SOON AS POSSIBLE AFTER COLLECTION
IS COMPLETED BECAUSE THE EXTRACTION HOLD-TIME FOR THESE SAMPLES
IS 7 DAYS AFTER COLLECTION.

THE SAMPLES MUST REMAIN COLD SO ALL  SHIPMENTS SHOULD BE MADE
MONDAY THROUGH THURSDAY. BECAUSE  SAMPLES ARE NOT DELIVERED ON
THE WEEKEND, NO SHIPMENTS SHOULD BE MADE ON FRIDAY.
V.   PREVENTATIVE MAINTENANCE

     A.   Schedule

         1.    Weekly

              a.    Check power cords and plugs, repair or replace as necessary.

              b.    Check filter holder gasket for leakage, replace if worn.

              c.    Check shelter hinges for breaks or stiffness, repair if necessary.

              d.    Check tubing from Flow Venturi to Magnehelic gage for holes
                   and/or cracking.

VI.   DATA FORMS

     All sample related run data forms will be supplied by ERG. Check the data sheets
     for completion after every setup or retrieval event.  The operator is expected to
     keep a logbook to document all  site activities, quality assurance activities,  and
     sampling activities. The ERG  PM10 Metals Sample Data Sheet is attached below.
                               Page 14 of 15

-------
                                                 School Air Toxics, SVOC / PAH SOP
                                                                   July 30, 2009
           Figure 3 SVOC Chain of Custody (COC)
                                            E^GLablDff
                    SVOC SAMPLE DATA SHEET

eld Sell) p
u
£
5
      Site Cede
                                    ColecUcri ^ate:
      AJGG Cnce
      Cartrtigt Geri'cjcoi Dacs:
site i


Co ecflcri ;ste:
                                          Sy&ten*:
                                                    RsB3t iV'N;:
                          Coll&cton Sys:6m Information:

S:at
End
EJapcpd T1IT4


^y
Tampi^C>



BarotirtrlGC'Hol



MagnahallD
ITHjOl



ulc r3'mln;i



      Tola CD lector ~lm

      --j-^:     Vald
                                         Tulal ColecUor '.^um£<:itd.
                              ;C-rctecr:
       tie^td by:
              Valfl
                              i;Crcfccre:
                                         ^efflne-Htw 4o

                                           TEmserixri:
      U VD C,
                            Page 15 of 15

-------