September 2009
    Environmental Technology
    Verification Report
    ABRAXIS ECOLOGENIA 17p-EsTRADioi_
    (E2) MlCROPLATE ENZYME-LINKED
    IMMUNOSORBENT ASSAY (ELISA) TEST KITS
              Prepared by
               Battelle

              Bafteiie
             1ne Business of Innovation
          Under a cooperative agreement with


            U.S. Environmental Protection Agency
ET1/ET1/ET1/

-------

-------
                                      September 2009
Environmental Technology Verification
                  Report

   ETV Advanced Monitoring Systems Center

ABRAXIS ECOLOGENIA® 17p-EsTRADioi_ (E2)
  MlCROPLATE ENZYME-LINKED IMMUNOSORBENT
           ASSAY (ELISA) TEST KITS
                     By

       Stephanie Buehler, Zachary Willenberg, Amy Dindal, Battelle

      Eric Kleiner, Michelle Henderson, and John McKernan, U.S. EPA

-------
                                        Notice

 The U.S. Environmental Protection Agency,  through its Office of Research and Development,
funded and managed, or partially funded and collaborated in, the research described herein. It
 has been subjected to the Agency's peer and administrative review and has been approved for
publication. Any opinions  expressed in this report are  those of the author (s) and do not
 necessarily reflect the views of the Agency, therefore, no official endorsement should be inferred.
Any mention  of trade  names or  commercial products does not constitute endorsement or
 recommendation for use.
                                           11

-------
                                      Foreword

The EPA is charged by Congress with protecting the nation's air, water, and land resources.
Under a mandate of national environmental laws, the Agency strives to formulate and implement
actions leading to a compatible balance between human activities and the ability of natural
systems to support and nurture life. To meet this mandate, the EPA's Office of Research and
Development provides data and science support that can be used to solve environmental
problems and to build the scientific knowledge base needed to manage our ecological resources
wisely, to understand how pollutants affect our health, and to prevent or reduce environmental
risks.

The Environmental Technology Verification (ETV) Program has been established by the EPA to
verify the performance characteristics of innovative environmental technology across all media
and to report this objective information to permitters, buyers, and users of the technology, thus
substantially accelerating the entrance of new environmental technologies into the marketplace.
Verification organizations oversee and report verification activities based on testing and quality
assurance protocols  developed with input from major stakeholders and customer groups
associated with the technology area. ETV consists of six environmental technology centers.
Information about each of these centers can be found on the Internet at http://www.epa.gov/etv/.

Effective verifications of monitoring technologies are needed to assess environmental quality
and to supply cost and performance data to select the most appropriate technology for that
assessment. Under a cooperative agreement, Battelle has received EPA funding to plan,
coordinate, and conduct such verification tests for "Advanced Monitoring Systems for Air,
Water, and Soil" and report the results to the community at large. Information concerning this
specific environmental technology area can be found on the Internet at
http ://www. epa.gov/etv/centers/centerl .html.
                                           in

-------
                                 Acknowledgments

The authors wish to acknowledge the support of all those who helped plan and conduct the
verification test, analyze the data, and prepare this report. We sincerely appreciate the
involvement and support of all staff from the participating laboratories who conducted testing as
part of this verification test. In particular, we would like to thank the following staff for their
contribution  in conducting this verification at their respective laboratories: Mark Mills and Scott
Jacobs, U.S.  EPANRMRL, Cincinnati, OH; Jennifer Gundersen, Dave Russell, Ronald Landy,
Annie Hilliard, John Curry, and Martin Lazarus, U.S. EPA Region 3 Fort Meade, MD; Dennis
Wesolowski, Larry Zintek, and Charles Steiner, U.S. EPA Region 5 Chicago, IL; Mike Meyer,
Keith Loftin, Larry Barber, and James Gray, USGS, Kansas; Jim Lazorchak, Tirumuru Reddy,
and Dan Bender, U.S. EPANERL, Cincinnati, OH; and Jeanette Van Emon, U.S. EPA NERL
Las Vegas, NV. Finally, we would like to thank Lisa Olsen, USGS; Paul Pennington, NOAA,
and Marion Kelly, U.S. EPA,  for their review of this verification report.
                                          IV

-------
                                      Contents
                                                                                  Page

Notice	ii

Foreword	iii

Acknowledgments	iv

List of Abbreviations	vii

Chapter 1 Background	1

Chapter 2 Technology Description	2

Chapter 3 Test Design and Procedures	4
  3.1 Introduction	4
  3.2 Test Facilities	5
  3.3 Test Procedures	6
       3.3.1     Test Sample Collection and Preparation	7
       3.3.2     Test Sample Analysis Procedure	8

Chapter 4 Quality Assurance/Quality Control	10
  4.1     Quality Control Samples	10
       4.1.1 GC-MS Blank and Surrogate Spike Results	10
       4.1.2 Method Blanks	11
  4.2 Audits	11
       4.2.1 Performance Evaluation Audit	11
       4.2.2 Technical Systems Audit	12
       4.2.3 Data Quality Audit	13
  4.3 QA/QC Reporting	13
  4.4 Data Review	13

Chapters Statistical Methods	15
  5.1 Precision	15
  5.2 Percent Bias	16
  5.3 Matrix Effects	16
  5.4 Operational Factors	17

Chapter 6 Test Results	18
  6.1 Precision	18
  6.2 Percent Bias	19
  6.3 Matrix Effects	20
  6.5 Operational Factors	22

Chapter 7 Performance Summary	24

Chapters References	26

-------
                                       Figures

Figure 2-1. Abraxis 17|3-Estradiol (E2) microplate ELIS A Test Kit	2



                                        Tables


Table 3-1. Target Analytes	5

Table 3-2. ELISA Test Kit Evaluation Responsibilities for Each Participating Laboratory	5

Table 4-1. PE Audit Sample Results	12

Table 4-2. Summary of Data Recording Process	14

Table 6-1. ELISA Test Kit Average Concentration and Relative Standard Deviation (RSD)
Results	19

Table 6-2. ELISA Test Kit Percent Bias vs. GC-MS	19

Table 6-3. GC-MS Average Concentration, RSD, and Percent Bias Results	20

Table 6-4. ELISA Test Kit Percent Bias vs. Expected Spike Concentration	21

Table 6-5. Percent Recovery	21

Table 6-6. Concentrations of Cross-Reactive Compounds	22
                                          VI

-------
                             List of Abbreviations
AMS
°C
coc
DI
El
E2
E3
EDC
EE2
ELISA
EPA
ETV
GC-MS
OFF
HPLC
L
LC-MS
MB
MDL
NERL
ng
nm
NP
NRMRL
ORD
ppb
PE
QA
QC
QMP
rpm
RSD
S
Advanced Monitoring Systems
degrees Celsius
chain of custody
deionized
estrone
17-p-estradiol
estriol
endocrine-disrupting compound
17-a-ethynylestradiol
enzyme-linked immunosorbent assay
U.S. Environmental Protection Agency
Environmental Technology Verification
gas chromatography-mass spectrometry
glass fiber filter
high performance liquid chromatography
liter
liquid chromatography-mass spectrometry
method blank
method detection limit
microliter
micron
EPA ORD National Exposure Research Laboratory
nanogram
nanometer
nonylphenol
EPA ORD National Risk Management Research Laboratory
EPA Office of Research and Development
parts per billion
performance  evaluation
quality assurance
quality control
quality management plan
revolutions per minute
relative standard deviation
standard deviation
                                       vn

-------
SOP             standard operating procedure
SPE             solid phase extraction
TSA             technical systems audit
USGS            United States Geological Survey
v/v              volume/volume
WWTP          wastewater treatment plant
                                        Vlll

-------
                                      Chapter 1
                                     Background


The U.S. Environmental Protection Agency (EPA) supports the Environmental Technology
Verification Program (ETV) to facilitate the deployment of innovative environmental
technologies through performance verification and dissemination of information. The goal of
ETV is to further environmental protection by accelerating the acceptance and use of improved
and cost-effective technologies. ETV seeks to achieve this goal by providing high-quality, peer-
reviewed data on technology performance to those involved in the design, distribution, financing,
permitting, purchase, and use of environmental technologies.

ETV works in partnership with recognized testing organizations; with stakeholder groups
consisting of buyers, vendor organizations, and permitters; and with the full participation of
individual technology developers. The Program evaluates the performance of innovative
technologies by developing test plans that are responsive to the needs of stakeholders,
conducting field or laboratory tests (as appropriate), collecting and analyzing data, and preparing
peer-reviewed reports. All evaluations are conducted in accordance with rigorous quality
assurance (QA) protocols to ensure that data of known and adequate quality are generated and
that the results are defensible.

 The EPA's National Risk Management Research Laboratory (NRMRL) and its verification
organization partner, Battelle, operate the Advanced Monitoring Systems (AMS) Center under
ETV.  The AMS Center recently evaluated the performance of the Abraxis 17|3-estradiol (E2)
microplate enzyme-linked immunosorbent assay (ELISA) test kit for determining endocrine-
disrupting compounds (EDCs) in water.

-------
                                       Chapter 2
                               Technology Description
The objective of the ETV AMS Center is to verify the performance characteristics of
environmental monitoring technologies for air, water, and soil. This report provides results for
the verification testing of the Abraxis 17|3-estradiol (E2) microplate ELISA test kit. The
following is a description of the test kit, based on information provided by the vendor.
The 17|3-estradiol (E2) microplate ELISA kit applies the principle of ELISA to determine 170-
estradiol in water samples.  The E2 microplate ELISA kit uses a colorimetric procedure to detect
17|3-estradiol.

The standards and samples, and an enzyme labeled estradiol conjugate are added to a disposable
microtiter plate (uncoated), and mixed. 100 microliter (|jL) aliquots of the mixture are then
added to antibody (monoclonal anti-estradiol) coated wells in a 96-well microplate. At this point
a competitive reaction occurs between the estradiol which may be in the sample and the enzyme
labeled estradiol for a finite number of antibody binding sites.  The reaction is allowed to
continue for sixty (60) minutes. At the end of the incubation period, the plate contents are
decanted to remove the unbound reagents from the estradiol and labeled estradiol that remain
bound  to the antibodies on the  plate, in proportion to their original concentration. After
decanting, the plate is washed with Washing Solution.  A substrate is then added and
enzymatically converted from a colorless to a blue solution. After an incubation period, the
reaction is stopped by the addition of diluted acid. The estradiol concentration is determined by
measuring the absorbance of the sample solution with a photometer (450 nm) and comparing it
                                             to the absorbance of standards.
                                             The E2 microplate ELISA Kit (Figure 2-1)
                                             contains a 96-well microplate coated with
                                             estradiol antibody (mouse anti-estradiol), two
                                             vials of lyophilized horseradish peroxidase-
                                             labeled estradiol analog, two 7  mL vials of
                                             buffer solution, five 1.5 mL vials of estradiol
                                             standard concentrations of 0, 0.05, 0.15, 0.4, 1.0
                                             parts per billion (ppb) with preservatives and
                                             stabilizers, a 15 mL bottle of hydrogen peroxide
                                             and 3,3',5,5'-tetramethylbenzidine solution in an
                                             organic base, a 15 mL bottle of diluted acid, a
                                             50 mL bottle of 6X concentrated Washing
                                             Buffer, one uncoated microtiter plate, and one
                                             adhesive plate cover.
Figure 2-1. Abraxis 17p-Estradiol (E2) microplate
ELISA Test Kit

-------
The E2 microplate ELISA Kit measures 7 by 5 by 4 Va inches.  Final results and calibration
curves are printed from the photometric analyzer or sent directly to a laboratory computer. List
price is $699 for a 96-test kit. Other materials that are required but are not provided with the E2
microplate ELISA Kit are pipettes, and a plate photometer capable of reading at 450 nanometer
(nm). These materials can be purchased separately or rented.

-------
                                      Chapter 3
                            Test Design and Procedures
3.1 Introduction

This verification test was conducted according to procedures specified in the Test/QA Plan for
Verification of Enzyme-Linked Immunosorbent Assay (ELISA) Test Kits for the Quantitative
Determination of Endocrine Disrupting Compounds (EDCs) in Aqueous Phase Samples1
Deviations to the test/QA plan were made due to unanticipated circumstances. As such, the test
procedures described in this chapter are a complete description of the actual test conditions.

Because of their potential to interfere with human, domestic animal, and wildlife reproduction,
EDCs are of increasing concern throughout the country. Several EPA Regions have undertaken
activities to monitor for these compounds, and several states are considering including
monitoring for EDCs in their regulatory programs. Presently, gas chromatography-mass
spectrometry (GC-MS), high performance liquid chromatography (HPLC), and liquid
chromatography-mass spectrometry (LC-MS) are being used for detecting these compounds.
However, immunoassay techniques, particularly ELISA, are becoming increasingly popular in
the field of environmental analysis due to their high sensitivity, ease of use, short analysis time,
and cost-effectiveness.

Immunoassay analytical detection is based on the capability of antibodies to specifically
recognize and form stable complexes with antigens. Immunoassays employ antibodies as
analytical reagents. In ELISA test kits, an enzyme conjugate competes with the chemical in the
sample for a limited number of binding sites on the antibody coated plate or particle. The extent
of color development is inversely proportional to the amount of chemical in the sample or
standard. The higher the concentration of a specific steroid or other EDC in the sample, the less
color reaction produced and recorded using a plate reader or tube photometer.

Testing was conducted with multiple collaborating laboratories, specifically the EPA Office of
Research and Development (ORD) National Risk Management Research Laboratory (NRMRL),
EPA Region 3, and the United States Geological Survey (USGS) Organic Geochemistry
Research Laboratory in Kansas. The laboratory participation was coordinated by EPA NRMRL,
in collaboration with Battelle. Laboratory names are removed, and simply stated as "Laboratory
(or Lab) 1, 2,  and 3" in the test results section, since inter-laboratory comparison was not an
objective of this report.

-------
This verification test evaluated the ability of the Abraxis E2 microplate ELISA test kit to
quantitate 17-p-estradiol (E2) in four different water matrices, per the manufacturer protocols.
More detailed information on the EDC tested  is provided in Table 3-1. EPA and USGS
laboratories used the Abraxis E2 microplate ELISA kit (according to Table 3-2) to quantitate
triplicate spiked  samples for hormones (E2), which were prepared and shipped by EPA NRMRL.
The test was conducted in four phases from June to September 2008, with each phase being a
different aqueous matrix.  As the more established method for detecting these compounds, GC-
MS served as the reference method2 for this test.

Table 3-1. Target Analytes
Analyte
(l?p)-Estra-l,3,5(10)-
triene-3,17-diol
Synonyms
17-p-Estradiol (E2)
CAS#
50-28-2
Use
Naturally occurring
hormone
Table 3-2. ELISA Test Kit Evaluation Responsibilities for Each Participating Laboratory
Responsibility
Sample Collection, Processing and Distribution
Test Kit Evaluation - E2 96 well ELISA
Reference Measurement - E2 GC-MS
NRMRL
A/
A/
A/
Region 3

A/

USGS-KS

A/

The Abraxis E2 microplate ELISA test kit was verified by evaluating the following parameters:

•  Precision
•  Percent bias
•  Matrix effects
•  Operational factors.

Verification of the system was conducted from June to September 2008. Precision was
determined by measuring the relative standard deviation of average concentration values as
reported by the test kit. Percent bias was determined as positive or negative, with positive values
indicating that ELISA concentration was higher than the reference method and negative values
indicating that it was lower. Matrix effects were determined by comparing the percent bias
measurements for Phase I deionized (DI) water samples to the percent bias measurements for the
Phase II through IV matrix-water samples.

Operational factors were determined based on documented observations of the testing staff and
the Verification Test Coordinator. Operational factors were described qualitatively, not
quantitatively; therefore, no statistical approaches were applied to the operational factors.
3.2 Test Facilities

Laboratory analyses of the Abraxis E2 microplate ELISA test kit were conducted in three
different collaborating laboratories by the laboratory staff.  These laboratories were: EPA ORD
NRMRL laboratory in Cincinnati,  OH; EPA Region 3; and USGS - Kansas.
                                           5

-------
3.3 Test Procedures

This verification test was conducted in four phases. Phase I consisted of a clean water sample
(DI water) spiked with a single concentration of E2, split into single samples, and submitted to
the ELIS A kit users in each collaborating laboratory to measure the concentration, in triplicate.
The split sample, as well as the un-spiked, matrix background sample, were also simultaneously
sent for reference GC-MS analysis of E2 and for various compounds which are known to cause
cross-reactivity with some ELIS A kits.  Phase II consisted of environmental surface water
samples subjected to the same spiking and splitting process as Phase I. Phase III consisted of a
complex matrix of wastewater treatment plant (WWTP) effluent samples subjected to the same
spiking and splitting process as Phase I and II. Phase IV consisted of a complex matrix of
WWTP influent samples, spiked and split as in previous phases. Details on the sample matrices,
spiking levels, and spiking procedures for each Phase are provided in Section 3.3.1. All E2 spike
concentrations used in each phase of this verification test were based on real-world
concentrations found in environmental samples, per the procedure described in the test/QA plan.1
Background concentrations of E2 were taken for each matrix for each phase. These GC-MS
measurements were made to determine if any measureable amounts of E2 might exist in the
sample matrix prior to the addition of any sample spikes. If a detectable concentration was
found, this concentration was then added to the spiked amount of E2 to calculate the total
concentration for all spiked samples of a particular phase.  Specific concentrations of E2, as
presented in Section 3.3.1, were spiked into the sample matrix for each phase, regardless of any
background concentrations of these compounds that may have been present in the collected
water. For Phase III and Phase IV, 4.03  ng/L and 4.00 ng/L of E2, respectively, were found in
the background matrix samples.  The nominal concentration of each sample was then calculated
using the measured background concentration and the expected spiked concentrations for each
phase. The ELISA kit results from the various laboratories for the Abraxis test kit were
compared to each other and compared to GC-MS results.

The E2 microplate ELISA test kit was tested only under laboratory controlled conditions, as
opposed to field conditions which would have been more variable. The analyses were performed
according to the vendor's recommended procedures as described in the user's instructions or
manual.  Simple cleanup procedures, as directed by the manufacturer of the test kit, were used
for the four different matrices. Each sample was analyzed after solid phase extraction (SPE)
cleanup using the procedure detailed in the kit instructions and provided in Section 3.3.2. Each
sample for ELISA analysis was filtered through a 1 micron (um) glass fiber filter prior to SPE.
Calibration and maintenance of the technology reader (i.e., microplate reader) was performed as
specified by the vendor.

A US EPA NRMRL GC-MS standard operating procedure (SOP) was followed for reference
measurements.  The GC-MS method for estrone (El), E2, estriol  (E3), EE2, testosterone,
dihydrotestosterone, androstenedione, and progesterone operated within a concentration range of
2-50 ng/L.  Samples for the GC-MS methods went through an extraction step to concentrate (or
dilute, depending upon the sample) to ensure the samples were within the method's analytical
range.2 The procedures for preparing, storing, and analyzing the test samples are provided
below.

-------
3.3.1   Test Sample Collection and Preparation

All sample bottles and glassware associated with hormone samples, including the glass carboy,
were cleaned and silanized using a procedure included in the test/QA plan1. All samples were
thoroughly mixed and were thus assumed to contain the same concentration. Samples were
spiked with E2 as one large stock solution and then split into smaller sub-samples in bottles. All
sample bottles were amber glass to prevent photodegradation of the analytes. All samples were
prepared and shipped by NRMRL, immediately after being made, in coolers on ice or freezer
packs to maintain a 4 degrees Celsius (°C) temperature.  When samples were received by each
laboratory, the condition of the samples, i.e., temperature, broken bottles etc., was noted by the
receiving laboratory operator and the samples were then immediately placed in a refrigerator at
4°C until analyzed. Holding times of hormone samples are currently unknown; therefore, all
samples were either analyzedor solvent exchanged within 24 hours of receipt to reduce error
associated with analyte degradation during sample holding. All laboratories performing
quantitative analysis, ELISA or GC-MS, received split samples from the same bulk sample.
Each laboratory that participated in the ELISA analysis received one 2.5 L spiked sample plus
one 500 mL DI water method blank.  The laboratory that performed the reference analysis
received one 4 L spiked sample and one 1 L DI water method blank to be processed by the GC-
MS method.

3.3.1.1 Phase I Samples

A sample of DI water was collected in a cleaned, 20 L, glass carboy from the USEPA laboratory
in Cincinnati, Ohio. The water was spiked with E2 to a concentration of 10 ng/L. This
concentration was selected because they are on the higher end of the range of concentrations
expected to be encountered in a real-world situation and is representative of the anticipated mid-
range of the test kit. The carboy was thoroughly mixed, by inserting a stir bar  and stirring on a
stir plate at 300 revolutions per minute (rpm) for 2 hours, to ensure homogeneous concentrations
of the analyte throughout the carboy.  One 2.5 L spiked sample was collected for each
participating laboratory as well as one 4L sample for each reference laboratory. DI water blanks
were also prepared and shipped in separate 500 mL bottles. The blank samples were analyzed
after SPE but only in two wells on the kits as opposed to three wells for all other  samples.
Before spiking, the DI water was sampled and analyzed by GC-MS to confirm the background
levels of E2.  Samples of the spiked mixtures were taken and the  concentrations of these  samples
and blank were determined using the E2 microplate ELISA test kit and GC-MS.

3.3.1.2 Phase II Samples

Grab samples of stream water were collected in three, clean, five gallon buckets from the South
Hasha Tributary to Eastfork Lake in Clermont County, Ohio. The tributary was accessed from
where it crosses Williamsburg-Bantam Road.  Before the stream water was spiked, a single
sample of the collected stream water was taken, split into triplicate aliquots,  and analyzed by
GC-MS to confirm the background levels of E2. Background levels of E2 measured in the
samples were added to the spiked concentration of E2 once results were obtained. Next,  a
cleaned, 20 L, glass carboy was used to collect 20 L of the stream water, which was then spiked
to contain a 10 ng/L concentration of E2. The carboy was thoroughly mixed by inserting a stir
bar and stirring on a stir plate at 300 rpm for 2 hours, to ensure homogeneous concentration of
the analyte throughout the carboy. Split samples were taken, as noted in Phase I. DI water

                                           7

-------
method blanks were filled with DI water at the same time as the stream water in the 20 L
carboys.

3.3.1.3 Phase III Samples


Grab samples of final effluent wastewater were collected in three, clean, five gallon buckets from
the Metropolitan Sewer District of Greater Cincinnati in Hamilton County, Ohio. After the
sample was transported back to the NRMRL laboratory, the effluent was measured and then
transferred into a clean, 20 L carboy. Before spiking, a single sample of the effluent was taken,
split into triplicate aliquots, and analyzed by GC-MS to confirm the background levels of E2. In
a cleaned,  20 L, glass carboy, 20 L of WWTP effluent was prepared containing 10 ng/L of E2.
The carboy was thoroughly mixed by inserting a stir bar and stirring on a stir plate at 300 rpm for
2 hours, to ensure homogeneous  concentration of the analyte throughout the carboy.   Split
samples were collected, as noted for Phase I.

3.3.1.4 Phase IV Samples

Grab samples of influent wastewater were collected in three, clean, five gallon bucket from the
Metropolitan Sewer District of Greater Cincinnati in Hamilton County, Ohio. After the sample
was transported back to the NRMRL laboratory, the influent was measured and transferred into a
20 L carboy. Before spiking, a single sample of the influent was taken, split into triplicate
aliquots, and analyzed by GC-MS to confirm the background levels of E2. In a cleaned, 20 L,
glass carboy, 20 L of WWTP influent was prepared containing 10 ng/L concentration of E2. The
carboy was thoroughly mixed by inserting a stir bar and stirring on a stir plate at 300 rpm for 2
hours, to ensure homogeneous concentration of the analyte throughout the carboy. Split samples
were collected as noted in Phase I.
3.3.2   Test Sample Analysis Procedure

The ELISA test kit users followed simple cleanup procedures as directed in the vendor's
instructions. The 2.5L sample was split into three 500 mL aliquots. Each of the three aliquots
was analyzed by utilizing glass fiber filter (GFF) cleanup and SPE. Each aliquot sample was
transferred in triplicate to the 96-well microplate for quantification, per the test kit protocols.
The cleanup procedures are described below.

Each sample for ELISA analysis was filtered through a 1 |Jm GFF.  Three 500 mL aliquots were
removed from the filtered sample for SPE. These three aliquots were treated as three
independent samples. SPE directions entitled "Extractions for EE2 from Water Sample for
ELISA", which were based on the vendor's protocols and summarized by EPA NRMRL, were
followed.1  The SPE protocol consists of the following steps:
1. Filter 500 mL of the sample, or the remainder of liquid in the sample bottle noting the volume
for later calculation, through 1 |j,m glass fiber filter.
2. Rinse a CIS SPE cartridge with 5mL of methanol and then 10 mL of distilled water at a flow
rate not exceeding 20 mL/min (preconditioning).
3. Pour the filtered sample through the CIS  SPE cartridge at a flow rate, no faster than 20
mL/min.

-------
4. Wash the cartridge with 5 mL of distilled water (up to 20 mL/min). Keep suctioning for about
a minute to dry the cartridge.
5. Wash the cartridge with 5 mL of hexane (up to 20 mL/min).
6. Elute the analyte with 5 mL of dichloromethane at a rate, no faster than 3 mL/min.
7. Evaporate the solvent with nitrogen gas to dryness.
8. Add 1 mL of 100% methanol to the residue and stir the mixture with a vortex mixer.  To
adjust the content to 10% methanol (volume/volume (v/v)) add 9 mL of distilled water for a total
volume of 10 mL.

After the SPE column, the E2 samples were reconstituted with 10 mL of a 10% methanol
solution.  For the spiked samples,  this process effectively increased the overall spike
concentration 50 times to an expected level of 505 ng/L. All reconstituted samples were applied
to three wells (100  |jL for each well) according to the manufacturer's instructions. Samples
were quantified by  reading their photometric responses  at a wavelength of 450 nm using a plate
reader following the manufacturer's instructions. The general steps for operating the Abraxis E2
microplate ELISA test kit that were followed during this verification test  are provided below.

The Abraxis ELISA E2 96 well microtiter plate kit assay procedure consists of the following
steps:
1. Take the kit out of the refrigerator approximately half an hour before use and let come to room
temperature (18-25°C). Filter the sample through a lum glass fiber filter and add  methanol to
obtain a final methanol concentration of 10% (v/v).
2. Dilute the E2 standard concentrate 10 fold with 10%  methanol to prepare a 10 |j,g/L E2 diluted
stock solution. Then, further dilute this solution to prepare E2 standards of designated
concentrations, from 50 ng/L to 1000 ng/L.
3. Reconstitute the  antigen-enzyme conjugate powder with 7mL out of 8mL buffer solution.
4. Mix 100 |jL of conjugate solution  and  100 (jL of E2 standard (or sample) in each well of
uncoated plate. Dispense the conjugate solution first, then add standard solution of sample.
Each standard will  be added to  two wells while each sample will be added to four  wells.
5. Dispense 100 uL aliquots of the standard/sample and conjugate mixture into each coated well
of the microplate.  Incubate it for 60  minutes at room temperature (18-25°C).
6. Dilute wash solution (6-fold  concentration) in 5 times of its volume of distilled  water to
prepare a wash solution.
7. Rinse each microplate well with approximately 300 uL of the wash solution and repeat the
step twice more. Firmly tap out the plate on a lint-free paper towel to remove solution from the
microplate.
8. Mix chromogen  solution and substrate solution at a ratio of 1:100 to prepare coloring reagent.
9. Dispense 100 uL of the color solution into each well  and incubate it for 30 minutes at room
temperature (18-25°C). Then, add 100 uL of stop solution to terminate reaction.
10. Measure the absorbance at 450 nm for each standard solution and generate a standard curve.
The quantity of E2  in the sample is then calculated from an absorbance reading and interpolation
from the standard curve.

-------
                                      Chapter 4
                        Quality Assurance/Quality Control


QA/Quality Control (QC) procedures were performed in accordance with the quality
management plan (QMP) for the AMS Center3 and the test/QA plan for this verification test.1
Test procedures were as stated in the test/QA plan1; however deviations to the test/QA plan were
made due to unanticipated circumstances. As such, the test procedures described in Chapter 3 are
a complete description of the actual test conditions. The statistical calculations intended for
analysis of the test kit results were also changed. This deviation is further described in Chapter
5. This change had no impact on the quality of the results. QA/QC procedures and results are
described below.
4.1 Quality Control Samples

Steps taken to maintain the quality of data collected during this verification test included
analyzing specific quality control samples for both the reference method (GC-MS) and the test
kits.

4.1.1 GC-MS Method Blank and Surrogate Spike Results

This verification test included a comparison of the Abraxis E2 microplate ELISA test kit results
to those of the GC-MS reference method for E2. Samples analyzed for each phase included PE
samples, test samples, background samples, and blank samples.  The quality of the reference
measurements was evaluated by adherence to the requirements of the GC-MS method for this
compound,  including requirements for method blanks (MBs), instrument solvent blanks, and
surrogate spikes, as indicated in the test/QA plan.l Method blank samples were analyzed to
ensure that no sources of contamination were present. If the analysis of a method blank sample
indicated a concentration above five times the method detection limit, contamination was
suspected. Any contamination source(s) were corrected and samples were reanalyzed or flagged
before proceeding with the analyses.  Surrogate spikes were also included in each sample.
Average acceptable recoveries for these samples were between 60 and 140%.  Samples outside of
the acceptable range were generally flagged and/or reanalyzed. D4-EE2 was used as a  surrogate
standard for the GC-MS analysis of E2 in the samples.  No levels of E2 were detected in any of
the reference method blank samples.

Surrogate recoveries in Phase I - IV samples varied across phases. Phase I surrogate recoveries
ranged from 59 - 96% and  averaged 85 ± 10% across 13 samples.  All recoveries were
considered in the acceptable range. Phase II recoveries ranged from 77 to  155% and averaged

                                           10

-------
132 ± 25% over 11 samples.  Surrogate recoveries for six of the samples were outside of the
acceptable range.  Compared to the surrogates, the peak shapes for the target analytes were good
and the baselines were clean in the chromatogram. Phase III surrogate recoveries ranged from
154% to 197% and averaged  176 ± 14% over 8 samples.  Surrogate recoveries for all samples
were outside of the acceptable range for Phase III. Phase IV surrogate recoveries ranged from
61% to 93% and averaged 76 ± 10% over 11 samples.  Surrogate recoveries for all Phase IV
samples were within the acceptable range.
4.1.2  Test Kit Method Blanks

Method blank samples were run in duplicate after SPE clean-up with each set of samples for all
four phases. Method blank samples were unspiked DI water.  Because concentrations for
samples analyzed with the test kit are calculated based on the interpolations from a curve
constructed from the standards run with each batch of samples, it is possible to obtain
concentration values for all samples. However, the E2 microplate test kit has a stated method
detection limit (MDL) of 0.05 - 1 [ig/L.  Based on this MDL,  it is assumed that sample
concentrations lower than this level  cannot be reliably determined or reported. Thus, any
samples, including method blank samples, with concentrations lower than the manufacturer's
stated MDL were considered non-detects.

The E2 microplate test kit was evaluated by three laboratories (see Table 3-2). Concentrations of
E2 were not detected in any of the method blank samples from two of the participating
laboratories. For the USGS results,  levels of E2 above the MDL were not detected in Phase II -
IV method blanks. However, during the analysis of Phase I samples, concentrations of E2 above
the MDL were found for one of the two replicates in both sets of duplicate method blanks for one
test kit.  Two Abraxis E2 microplate ELISA test kits were evaluated by this laboratory for each
phase of testing.  The second test kit operated during Phase I did not show any detectable levels
of E2 in any of the method blanks analyzed. The detectable levels of E2 found in the method
blanks were only slightly above the MDL threshold, and averaging the  duplicate samples put the
overall concentration below the MDL. All method blanks for this phase came from the same
initial sample.
4.2 Audits

Three types of audits were performed during the verification test: a PE audit of the reference
method measurements (GC-MS analyses), a technical systems audit (TSA) of the verification
test performance, and a data quality audit. Audit procedures are described further below.

4.2.1  Performance Evaluation Audit

A PE audit was conducted to assess the quality of the reference method measurements (GC-MS
analyses) made in this verification test. The reference method PE audit was performed by
supplying an independent second standard solution of E2 prepared from a different source other
than that used in verification testing. The PE audit samples were analyzed in the same manner as
all other samples and the analytical results for the PE audit samples were compared to the
nominal concentration. The target criterion for this PE audit was agreement of the analytical
result within 30% of the expected concentration. This audit was performed once during  each
                                          11

-------
phase of testing.  Table 4-1 shows the percent error results for the PE samples for each phase.
The percent error was calculated based on the difference between the actual and expected E2
concentrations divided by the expected concentration.  The E2 PE audit samples were within
30% of the expected concentration for Phases I, II and IV, while Phase III PE audit samples were
outside of this agreement range (94% error). The PE results were consistent with the surrogate
results presented in Section 4.1.1, where Phases I and IV were within specifications; Phase II
was slightly outside, and Phase III did not meet requirements. The PE audit sample results were
also similar to the GC-MS sample results for Phase III.  This could have impacted the
comparison of the ELISA test kit results to the GC-MS data for Phase III. No adjustments were
made to the standards nor were PE audit samples reanalyzed based on these results. It does not
appear that the reference laboratory results for Phase II and III were overall of lower quality than
Phases I and IV,  since reference results were fairly consistent within phases for the actual
samples (e.g., see Table 6-2). However, the test kit results were more comparable to the expected
spiked concentrations (Table 6-3 and Table 6-4) than to the GC-MS results (Table 6-2).

Table 4-1. PE Audit Sample Results
Phase
1
II
III
IV
Expected
Concentration (ng/L)
10
10
10
10
Actual
Concentration (ng/L)
9.03
7.43
19.4
8.17
% Error
-10
-26
94
-18
4.2.2  Technical Systems Audit

The Battelle Quality Manager performed a TSA twice during this verification test. Because the
testing was taking place in multiple laboratories across the country, Battelle's Quality Manager
visited only two laboratories for in-person TSAs. Battelle conducted TSAs at the Cincinnati, OH
facility on July 23-24, 2008 and at the Fort Meade, MD facility on July 31, 2008. All TSA
findings were reported to the Verification Test Coordinator.

The purpose of this audit was to ensure that the verification test was being performed in
accordance with the AMS Center QMP,3 the test/QA plan,1 and the GC-MS SOP2 used during
the verification test. In the TSA, the Battelle Quality Manager reviewed the reference methods
used, compared actual test procedures to those specified or referenced in test/QA plan,  and
reviewed data acquisition and handling procedures. The Battelle Quality Manager also  toured the
laboratory where verification and reference testing were taking place,1 inspected sample chain of
custody (COC) documentation, reviewed technology-specific record books, checked standard
certifications and technology data acquisition procedures, and conferred with technical  staff. A
TSA report was prepared, including a statement of findings and the actions taken to address any
adverse findings, and a copy of Battelle's TSA report was sent to the EPA AMS Center QA
Manager. No adverse findings were reported. The TSA findings were communicated to
technical staff at the time of the audit.
                                           12

-------
4.2.3  Data Quality Audit

At least 10% of the data acquired during the verification test were audited. Battelle's Quality
Manager traced the data from the initial acquisition, through reduction and statistical analysis, to
final reporting to ensure the integrity of the reported results. All calculations performed on the
data undergoing the audit were checked. Minor transcription errors and errors due to rounding
were identified and corrected before the results were used for the calculations described in
Chapter 5.
4.3 QA/QC Reporting

Each audit was documented in accordance with Section 3.3.4 of the AMS Center QMP.3 Once
the audit reports were prepared, the Battelle Verification Test Coordinator ensured that a
response was provided for each adverse finding or potential problem and implemented any
necessary follow-up corrective action. The Battelle Quality Manager ensured that follow-up
corrective action was taken. The results of the TSA were submitted to the EPA.
4.4 Data Review

Records generated in the verification test received an independent internal review before these
records were used to calculate, evaluate, or report verification results. Table 4-2 summarizes the
types of data recorded. Data were reviewed by a Battelle technical staff member involved in the
verification test. The person performing the review added his/her initials and the date to a hard
copy of the record being reviewed.
                                            13

-------
Table 4-2. Summary of Data Recording Process
Data Recorded
Dates times and
details of test
events
Technology
calibration
information
Technology
readings
Sample collection
and reference
method analysis
procedures,
calibrations, etc.
Reference method
results
Where Recorded
Laboratory record
books or data
recording forms, or
electronically
Laboratory record
books, data
recording forms, or
electronically
Recorded
electronically or
manually by the
operator or
electronically by
the technology
reader, as
appropriate
Laboratory record
books, chain-of-
custody,
electronically, or
other data
recording forms
Electronically from
reference
measurement
technology
How Often
Recorded
Start/end of test
procedure, and at
each change of a
test parameter or
change of
technology status
At technology
reader calibration
or recalibration, as
applicable
Each sample and
QC analysis
Throughout
sampling and
analysis processes
Every sample or
QC analysis
By Whom
Participating
laboratories
Participating
laboratories
Participating
laboratories
Participating
laboratories
Participating
laboratories
Disposition of
Data
Used to organize
and check test
results; manually
incorporated into
data spreadsheets
as necessary
Incorporated into
verification report
as necessary
Converted to or
manually entered
into spreadsheets
for statistical
analysis or
comparisons
Retained as
documentation of
sample collection
or reference
method
performance
Transferred to
spreadsheets for
calculation of
results and
statistical analysis
or comparisons
                                       14

-------
                                       Chapter 5
                                  Statistical Methods
The statistical methods used to evaluate the quantitative performance factors listed in Section 3.1
are presented in this chapter. Qualitative observations were also used to evaluate verification test
data.

Per the test/QA plan,l repeatability and reproducibility were intended to be calculated as
performance parameters for this verification test.  However, after further discussion with EPA,
and in agreement with EPA, it was determined that higher level summary  statistics provided a
better synopsis of the test kit results. Thus, the mean and relative standard deviations (precision)
were calculated for the test kit results.
5.1 Precision

The standard deviation (S) of the results for the replicate analyses of the same sample was
calculated as follows:
                                        n-l.. .

where n is the number of replicate samples, M is the ELISA test kit measurement for the k
                                          K-
sample, and Mis the average ELISA test kit measurement of the replicate samples. The precision
for each sample is reported in terms of the relative standard deviation (RSD), which was
calculated as follows:
                                   RSD(%) =
                                              S
                                             M
xlOO
                                       (2)
The RSD was calculated for each laboratory that participated in the verification test and for each
test kit that was tested.  The RSD was also calculated across all laboratories and test kits for each
phase of testing.
                                           15

-------
5.2 Percent Bias

Percent bias was calculated as a percentage for each measurement in each phase for each kit
using Equation 3 :


                           %Bias = [ — -1]  xlOO                                  (3)
where y = 1, 2, 3 denotes the laboratory, / = 1, 2 denotes the ELISA test kit within laboratory, n =
1, 2 denotes the reference method, Xjt is the ELISA concentration for the/ laboratory and the i'
test kit, yn is the concentration of the reference method GC-MS or the concentration of the spike.
Ideally percent bias results will be within ±25%.
5.3 Matrix Effects

Matrix effects were examined by comparing the percent bias measurements for the Phase IDI
water samples to the percent bias measurements for the Phase II - IV samples. Percent bias was
determined as described in Section 5.2.

General observations of potential matrix effects, such as false negatives, if observed, were
documented but were not used in statistical calculations. False negatives were defined as a
negative (zero) response in a sample that is spiked with contaminant at a detectable
concentration.

General observations on potential cross-reactivity were documented. Blank samples of each
matrix were evaluated by GC-MS to determine background levels of the compounds with which
the kits have cross-reactivity, as stated by the vendor.

Percent recovery results were calculated on a per-sample and per-phase basis and were based on
the expected spiked concentration of the analyte in each sample matrix. Percent recovery was
calculated using the Equation 4:

                                         ^
                           % Recovery = — x 100                                     (4)
                                         E
Where A is the actual ELISA test kit measurement and E is the expected concentration.  The
expected concentration includes the known spike concentration as well as any detected
background levels of E2 in the matrix water (see Section 3.3). Percent recovery results are
presented to provide another measure of test kit performance to the end user. Ideal percent
recovery values are near 100%.
                                           16

-------
5.4 Operational Factors

Operational factors were determined based on documented observations of the testing staff.
Operational factors are described qualitatively, not quantitatively; therefore, no statistical
approaches were applied to the operational factors.
                                             17

-------
                                      Chapter 6
                                    Test Results
The results of the verification tests of the Abraxis E2 microplate ELISA test kit are presented
below for each of the performance parameters.
6.1 Precision

The relative standard deviation (RSD) is used as a means of evaluating the precision of the
ELISA test kit.  Three laboratories operated the E2 microplate ELISA test kit. Two laboratories
(Lab 1 and Lab 2) ran identical samples on two separate test kits (kit "a" and kit "b"). Lab 3 ran a
single kit. Table 6-1 presents the resulting RSD for each participating laboratory and test kit
along with the overall average concentrations per phase of E2 found using the E2 microplate
ELISA test kit for all analyses. RSD values are also presented across all results for each phase.

RSDs among the four phases ranged from 2 to 44%. The overall RSDs for the average
concentration of Phases II and IV were similar (18% and 16%, respectively). The Phase III
average concentration had the lowest RSD at 11%, close to the RSDs for Phases II and IV.
Phase I had the highest RSD among all analyses at 30%, which was approximately twice as high
as the RSD found among the remaining phases. The average concentration in Phase III was
higher than that of Phase IV, though both were expected to have the same concentration.
Similarly, the average concentration of Phase I was higher than that of Phase II.
                                          18

-------
Table 6-1. ELISA Test Kit Average Concentration and Relative Standard Deviation (RSD)
Results"
Phase I
Phase II
Phase
Phase IV

Lab 1 kit a
Lab 1 kit b
Lab 2 kit a
Lab 2 kit b
Lab 3
Lab 1 kit a
Lab 1 kit b
Lab 2 kit a
Lab 2 kit b
Lab 3
Lab 1 kit a
Lab 1 kit b
Lab 2 kit a
Lab 2 kit b
Lab 3
Lab 1 kit a
Lab 1 kit b
Lab 2 kit a
Lab 2 kit b
Lab 3
Average
Cone (ng/L)
597
595
443
408
655
458
463
543
387
596
857
940
949
916
Out of rangeb
890
880
883
889
931
Overall
Average
RSD Cone (ng/L) RSD
19%
25%
33% 540 30%
44%
9%
12%
5%
8% 489 18%
19%
7%
10%
6%
13% 925 11%
13%

3%
11%
5% 874 16%
4%
2%
Expected
Phase
Cone (ng/L)


505




505




706




705


 The average concentration and RSD are based on all replicates within the detectable range of the test kit.
 ' Test kit results were above the upper end of the test kit's range.
6.2 Percent Bias

Bias is a systematic error that causes measurements to err in one direction, either high or low.
For this section, percent bias was calculated relative to the GC-MS reference method results.  A
positive percent bias indicates that the ELISA test kit concentration is higher than the reference
method, while a negative percent bias indicates that the ELISA test kit concentrations are lower
than the reference method.  Table 6-2 presents the percent bias results.

Table 6-2. ELISA Test Kit Percent Bias vs. GC-MS

Lab 1 kit a
Lab 1 kit b
Lab 2 kit a
Lab 2 kit b
Lab 3
Phase I
25
25
-7
-14
37
Phase II
-28
-27
-15
-39
-7
Phase III
-7
2
3
-1
14
Phase IV
93
91
91
93
102
                                            19

-------
Phase I percent bias results were mixed, with some showing positive bias and some showing
negative. Results from both test kits for Lab  1 and Lab 3 results showed positive bias while
those for Lab 2 showed negative. Phase III samples were also mixed, with percent bias split
across test kits evaluated within a laboratory although the results indicated that the test kit and
GC-MS results for Phase III were similar since the percent bias values were small in magnitude.
All results for Phase II samples showed a negative bias while all Phase IV samples showed a
positive bias. Phase IV samples showed the highest bias consistently across all laboratories. All
biases in Phase IV were positive, covering a small range of positive bias from 91 - 102%. Lab 3
had the highest bias (either positive or negative) in most phases. Phase III had the least bias
across all laboratories, with three of the five tests having <5% (positive or negative) bias.

For comparison,  average concentrations, RSD, and percent bias for the GC-MS measurements
with regard to the expected concentration are presented in Table 6-3 for each phase. RSD values
were less than 30% for all  phases, and percent bias results were within ±35% of the expected
concentration. These results demonstrate that the GC-MS results were biased high (in Phases II
and III) and low  (in Phases I and IV) as compared to the expected concentration.

Table 6-3.  GC-MS Average Concentration, RSD, and Percent Bias Results
Phase
1
II
III
IV
Average
Cone (ng/L)
9.53
12.76
18.50
9.26
RSD
10%
3%
1%
7%
% Bias
(vs. Expected Cone)
-6
27
31
-34
6.3 Matrix Effects

To understand how the matrix of each phase of testing might have affected the results, percent
bias and percent recovery were calculated for the test kit results in comparison to the expected
spiked concentration of E2. A positive percent bias indicates that the ELISA test kit
concentration is higher than the expected spike concentration, while a negative percent bias
indicates that the ELISA test kit concentrations are lower than the expected spike concentration.
Table 6-4 presents the percent bias results. No false negatives were observed during this
verification test.

Phase III and Phase IV percent bias results were  all  positive. Though these phases generally
have the highest bias of all of the phases, all but  one of the results for these phases was below
35% bias. Phase I and II have a mix of positive and negative bias, indicating that the test kit
results erred both high and low as compared to the expected E2 spike concentrations. Lab 2 kit b
bias results were negative for both phases while kit  a results for Lab 2 were negative in Phase I
and positive in Phase II.  Regardless of whether the  bias was positive or negative, bias results for
all samples tested were similar within phases.  Lab 3 percent bias results were greater than
results from any other laboratory for Phases I, III, and IV.
                                           20

-------
Table 6-4. ELISA Test Kit Percent Bias vs. Expected Spike Concentration

Lab 1 kit a
Lab 1 kit b
Lab 2 kit a
Lab 2 kit b
Lab 3
Phase I
18
18
-12
-19
30
Phase II
-9
-8
8
-23
18
Phase III
21
33
34
30
>110
Phase IV
26
25
25
26
32
Phase III and Phase IV percent bias results were all positive. Though these phases generally
have the highest bias of all of the phases, all but one of the results for these phases was below
35% bias. Phase I and II have a mix of positive and negative bias, indicating that the test kit
results erred both high and low as compared to the expected E2 spike concentrations. Lab 2 kit b
bias results were negative for both phases while kit a results for Lab 2 were negative in Phase I
and positive in Phase II.  Regardless of whether the bias was positive or negative, bias results for
all samples tested were similar within phases. Lab 3 percent bias results were greater than
results from any other laboratory for Phases I, III, and IV.

In terms of positive or negative bias, the comparison of the test kit results to the expected spike
concentrations (Table 6-4) was similar to the comparison to the GC-MS results (Table 6-2). The
comparison to the expected concentrations usually showed a smaller amount of bias with the
exception of Phase III where the magnitudes were higher for the comparison to the expected
concentrations.

As another measure of accuracy, percent recovery results, comparing the test kit results against
the  expected spiked concentration, were also calculated on a per sample and per phase average
basis.  Table 6-5 presents these results.
Table 6-5. Percent Recovery

% Recovery
Lab 1 kit a
Lab 1 kit b
Lab 2 kit a
Lab 2 kit b
Lab 3
Average
Phase I
SPE
118
118
88
81
130
107
Phase II
SPE
91
92
108
77
118
97
Phase III
SPE
121
133
134
130
>149
130
Phase IV
SPE
126
125
125
126
132
127
Percent recoveries were close to 100% for Phases I and II and close to 130% for Phases III and
IV. Phase IV percent recoveries showed little variation between all testing laboratories.  Percent
recoveries were similarly close in Phase III for Labs 1 and 2. Of the 20 samples analyzed for all
phases, 15 were over-recovered. The percent recoveries for all phases were within the range of
acceptable recoveries for the GC-MS reference method.  Overall, the test kit results were more
comparable to the expected spiked concentrations (Table 6-4 and Table 6-5) than to the GC-MS
results (Table 6-2).
                                           21

-------
Some ELISA kits will react with compounds similar to the target compound, known as cross-
reactivity. The Abraxis E2 microplate ELISA E2 96 well plate test kit will react with known
percent reactivities to multiple hormones.  During each phase of the study, some of the
compounds with which the test kit has cross-reactivity were measured alongside background
levels of the kit's target compound in that matrix by GC-MS. For some of the compounds for
which there is known cross-reactivity with the E2 microplate ELISA test kit, there are no
established analytical methods available by GC-MS at these concentrations and in these
matrices. Therefore some error will have to be accepted from influence of cross-reactive
compounds that cannot be identified via GC-MS. According to the test kit instructions, there is
expected to be minimal error from these compounds compared to the primary target compound.
Table 6-6 lists concentrations found in each matrix  blank sample from each phase along with the
known percent reactivities for the cross-reactive compounds to the E2 microplate ELISA test kit
that were measured during this verification. Unfortunately, matrix blank samples used for
background analysis were not analyzed on the E2 microplate ELISA test kit.  Because of this, the
potential for cross-reactive compounds present in the matrix for each phase to interfere with the
test kit results cannot be truly evaluated.

Table 6-6.  Concentrations of Cross-Reactive Compounds
Steroid Hormones
Dihydrotestosterone
Estrone (E1)
Testosterone (TEST)
Estriol (E3)
Concentration (ng/L)
Phase I
ND
ND
ND
ND
Phase II
ND
1.4
ND
ND
Phase III
ND
44.9
1.7
1.4
Phase IV
43.7
16.8
23.2
11.6
Cross Reactivity (%)
0.38
1.3
<0.03
0.6
ND = not detected
6.5 Operational Factors

In general, training is needed to effectively and properly operate ELISA test kits. The vendor
trained staff on the operation of the test kit, but these trained staff were, in some cases, not
available for the verification test because of testing delays and staff turn over. Therefore, staff
that operated the test kits during the verification test may not have been trained by the vendor.

Operational concerns or issues were not reported from any of the three participating laboratories.
The test kit instructions were readily followed by each of the operators. Operation of the test kit
from the introduction of the sample until the reaction was stopped and the results were read, took
approximately 2-2.5 hours.  Preparation time was required prior to the introduction of the sample
to allow all reagents time to come to room temperature before using them. Calibrated pipettes,
test tubes, and a plate reader capable of reading at 450 nm are required for the operation of the
test kit, but are not supplied with the test kit. GFF or SPE equipment used with the samples were
                                           22

-------
not supplied with the test kit. For at least one laboratory, the concentration step after SPE was
time consuming, requiring 1-5 hours.

Each purchased test kit is capable of conducting 96 tests and costs $699. For comparison, GC-
MS analyses of these samples are estimated to cost between $500 and $900 per sample.l
                                          23

-------
                                      Chapter 7
                              Performance Summary
The ability of the Abraxis E2 microplate ELISA test kit to detect E2 in water was evaluated
using four different water matrices.  The test kit was operated by three different laboratories with
the use of SPE cleanup. The test kit results were evaluated against the expected spike
concentrations and the reference measurements of the same samples made using GC-MS.

Relative standard deviations (RSDs) among the four phases ranged  from 2 to 44%.  The overall
RSDs for the average concentration of Phases II and IV were similar (18% and 16%,
respectively). The Phase III average concentration had the lowest RSD at 11%, close to the
RSDs for Phases II and IV. Phase I had the highest RSD among all analyses at 30%, which was
approximately twice as high as the RSD found among the remaining phases. The average
concentration in Phase III was higher than that of Phase IV, though  both were expected to have
the same concentration. Similarly, the average concentration of Phase I was higher than that of
Phase II.

Percent bias, as compared to the GC-MS reference analysis results,  was mixed (positive and
negative) for Phases I and III.  All results for Phase II samples showed a negative bias while all
Phase IV samples  showed a positive bias. Phase III had the lowest percent bias across all
laboratories. Percent bias, as compared to the expected spiked E2 concentration, was positive for
Phase III and Phase IV. Phase I and II have a mix of positive and negative bias, indicating that
the test kit results erred both high and low as compared to the expected E2 spike concentrations.
Regardless of whether the bias was positive or negative, bias results for all samples tested were
similar within phases.

No false negatives were observed during this verification test. Average percent recoveries were
close to 100% for Phases I and II  and close to 130% for Phases III and IV. Of the 20 samples
analyzed for all phases, 15 were over-recovered.  The percent recoveries for all phases were
within the range of acceptable recoveries for the GC-MS reference method.

Operational concerns or issues were not reported from any of the three participating laboratories.
The test kit instructions were readily followed by each of the operators.  Operation of the test kit
from the introduction of the sample until the reaction was stopped and the results were read took
approximately 2-2.5 hours. Preparation time was required prior to the introduction of the sample
to allow all reagents time to come to room temperature before using them. Calibrated pipettes,
test tubes, and a plate reader capable of reading at 450 nm are required for the operation of the
test kit but are not supplied with the test kit. GFF and SPE equipment used with the samples was
                                          24

-------
not supplied with the test kit.  For at least one laboratory, the concentration step after SPE was
time consuming, requiring 1-5 hours. Each purchased test kit is capable of conducting 96 tests
and costs $699.
                                           25

-------
                                    Chapter 8
                                    References
1.   Test/QA Plan for Verification of Enzyme-Linked Immunosorbent Assay (ELISA) Test Kits
    for the Quantitative Determination of Endocrine Disrupting Compounds (EDCs) in Aqueous
    Phase Samples, Battelle, Columbus, Ohio, July 1, 2008.

2.   U.S. EPA NRMRL Standard Operating Procedure (SOP) for the Analysis of Steroid
    Hormones in Aqueous Samples, QA ID 503-P3-0, 09/29/05.

3.   Quality Management Plan (QMP)for the ETV Advanced Monitoring Systems Center,
    Version 6.0, U.S. EPA Environmental Technology Verification Program, Battelle,
    Columbus, Ohio, November 2005.
                                        26

-------