METHOD 556.1
   DETERMINATION OF CARBONYL COMPOUNDS IN
 DRINKING WATER BY FAST GAS CHROMATOGRAPHY
                             Revision 1.0

                            September 1999
J.W. Munch, US EPA Office of Research and Development, NERL
DJ. Munch, US EPA, Office of Ground Water and Drinking Water
S.D. Winslow, S.C. Wendelken, B.V. Pepich, ICF Kaiser Engineers, Inc.
Method 556, Revision 1.0 (1998)
S.C. Wendelken and B.V. Pepich (IT Corporation)
DJ. Munch, US EPA, Office of Ground Water and Drinking Water
Method 556.1, Revision 1.0 (1999)
              NATIONAL EXPOSURE RESEARCH LABORATORY
                OFFICE OF RESEARCH AND DEVELOPMENT
               U. S. ENVIRONMENTAL PROTECTION AGENCY
                       CINCINNATI, OHIO 45268

-------
                                   METHOD 556.1

DETERMINATION OF CARBONYL COMPOUNDS IN DRINKING WATER BY FAST
                             GAS CHROMATOGRAPHY
1.   SCOPE AND APPLICATION

    1.1    This is a fast gas chromatographic method optimized for the determination of
           selected carbonyl compounds in finished drinking water and raw source water. The
           analytes applicable to this method are derivatized to their corresponding
           pentafluorobenzyl oximes. The oxime derivatives are then extracted from the water
           with hexane. The hexane extracts are analyzed by fast gas chromatography with
           electron capture detection (FGC-ECD).  Fast GC uses a small diameter capillary
           column (< 100 |j,m i.d.), hydrogen carrier gas and a fast oven ramp rate to
           dramatically decrease analysis time.  Accuracy, precision, and method detection
           limit (MDL) data have been generated for the following compounds:

                                                   Chemical Abstract Services
           Analyte                                     Registry Number

           Formaldehyde                                     50-00-0
           Acetaldehyde                                     75-07-0
           Propanal                                          123-38-6
           Butanal                                           123-72-8
           Pentanal                                          110-62-3
           Hexanal                                          66-25-1
           Heptanal                                          111-71-7
           Octanal                                           124-13-0
           Nonanal                                          124-19-6
           Decanal                                          112-31-2
           Cyclohexanone                                   108-94-1
           Benzaldehyde                                     100-52-7
           Glyoxal (ethanedial)                               107-22-2
           Methyl glyoxal (2-oxopropanal or pyruvic  aldehyde)   78-98-8

    1.2    This method applies to the determination of target analytes over the concentration
           ranges typically found in drinking water. Analyte retention times are in Section 17,
           Tables 1 and 2.  Other method performance data are presented in Section 17, Tables
           3-7. Experimentally determined method detection limits (MDLs) for the above
           listed analytes are provided in Section 17, Table 4. The MDL is defined as the
           statistically calculated minimum amount that can be measured with 99% confidence
           that the reported value is greater than zero.(1'2) However, it should be noted that
           background levels of some method analytes (usually formaldehyde and
           acetaldehyde) are problematic.  The minimum reporting level (MRL) for method

-------
           analytes, for each analyst/laboratory that uses this method, will depend on their
           ability to control background levels (Section 4).

     1.3    This method is restricted to use by or under the supervision of analysts skilled in
           liquid-liquid extractions, derivatization procedures and the use of GC and
           interpretation of gas chromatograms. The analyst should be thoroughly familiar with
           the practice of fast GC before significantly modifying the conditions in Table 1.
           Each analyst must demonstrate the ability to generate acceptable results with this
           method, using the procedures described in Section 9.

2.   SUMMARY OF METHOD

     2.1    A 20 mL volume of water sample is adjusted to pH 4 with potassium hydrogen
           phthalate (KHP) and the analytes are derivatized at 35 °C for 2 hr with 15 mg of O-
           (2,3,4,5,6-pentafluorobenzyl)-hydroxylamine (PFBHA) reagent.  The oxime
           derivatives are extracted from the water with 4 mL hexane.  The extract is processed
           through an acidic wash step, and then analyzed by FGC-ECD.  The target analytes
           are identified and quantified by comparison to a procedural standard (Section 3.9).
           Two chromatographic peaks will be observed for many of the target analytes. Both
           (E) and (Z) isomers are formed for carbonyl compounds that are asymmetrical, and
           that are not sterically hindered. The (E) and (Z) isomers may not be
           chromatographically resolved in a few cases. Compounds with two carbonyl groups,
           such as glyoxal and methyl glyoxal, can produce even more isomers.  See Section
           17, Table 1 and Figure  1 for the chromatographic peaks used for analyte
           identification.

           NOTE:  The absolute  identity of the (E) and (Z) isomers was not determined during
                    method development. Other researchers (3A5) have reported the first
                    eluting peak as (E), and the second peak as (Z). For convenience, this
                    method will follow this convention. Because more than 2 isomers are
                    formed for glyoxal and methyl glyoxal, the peaks used for identification
                    are referred to as "peak 1" and "peak 2."

     2.2    All results should be confirmed on a second, dissimilar capillary GC column.

3.   DEFINITIONS

     3.1    LABORATORY REAGENT BLANK (LRB) -- An aliquot of reagent water or other
           blank matrix that is treated exactly as a sample including exposure to all glassware,
           equipment, solvents and reagents, sample preservatives, internal standards, and
           surrogates that are used with other samples.  The LRB is used to determine if
           method analytes or other interferences are present in the laboratory environment, the
           reagents, or the apparatus.

     3.2    FIELD REAGENT BLANK (FRB) - An aliquot of reagent water or other blank
           matrix that is placed in  a sample container in the laboratory and treated as a sample

-------
       in all respects, including shipment to the sampling site, storage, preservation, and all
       analytical procedures. The purpose of the FRB is to determine if method analytes or
       other interferences are introduced during sample shipping or storage. For this
       analysis the FRB should not be opened at the sampling site.

3.3     LABORATORY FORTIFIED BLANK (LFB) -- An aliquot of reagent water or other
       blank matrix to which known quantities of the method analytes are added in the
       laboratory. The LFB is analyzed exactly like a sample, and its purpose is to
       determine whether the methodology is in control, and whether the laboratory is
       capable of making accurate and precise measurements.

3.4     LABORATORY FORTIFIED SAMPLE MATRIX (LFM) -- An aliquot of an envi-
       ronmental sample to which known quantities of the method analytes are added in the
       laboratory. The LFM is analyzed exactly like a sample, and its purpose is to
       determine whether the sample matrix contributes bias to the analytical results.  The
       background concentrations of the analytes in the sample matrix must be determined
       in a separate aliquot and the measured values in the LFM corrected for background
       concentrations.

3.5     STOCK STANDARD SOLUTION (SSS) -- A concentrated  solution containing one
       or more method analytes prepared in the laboratory using assayed reference materials
       or purchased from a reputable commercial source.

3.6     PRIMARY DILUTION STANDARD SOLUTION (PDS) --  A solution of several
       analytes prepared in the laboratory from stock standard solutions and diluted as
       needed to prepare calibration solutions  and other needed analyte solutions.

3.7     CALIBRATION STANDARD (CAL) -- A solution prepared from the primary
       dilution standard solution and stock standard solutions and the internal standards and
       surrogate analytes. The CAL solutions are used to calibrate the instrument response
       with respect to analyte concentration.

3.8     QUALITY CONTROL SAMPLE (QCS)  -- A solution of method analytes of known
       concentrations which is used to fortify an aliquot of LRB or sample matrix. The
       QCS is obtained from a source external to the laboratory and different from the
       source of calibration standards. It is used to check laboratory performance with
       externally prepared test materials.

3.9     PROCEDURAL STANDARD CALIBRATION -  A calibration method where
       aqueous calibration standards are prepared and processed (e.g. purged, extracted,
       and/or derivatized) in exactly the same manner as a sample.  All steps in the process
       from addition  of sampling preservatives through instrumental analyses are included
       in the calibration. Using procedural standard calibration compensates for any
       inefficiencies in the processing procedure.

-------
    3.10   INTERNAL STANDARD (IS) - A pure analyte added to a sample, extract, or
           standard solution in known amount(s) and used to measure the relative responses of
           other method analytes and surrogates that are components of the same sample or
           solution.  The internal standard must be an analyte that is not a sample component.

    3.11   SURROGATE ANALYTE (SUR) - A pure analyte, which is extremely unlikely to
           be found in any sample, and which is added to  a sample aliquot in known amount(s)
           before extraction or other processing and is measured with the same procedures used
           to measure other sample components.   The purpose of the SUR is to monitor method
           performance with each sample.

    3.12   METHOD DETECTION LIMIT (MDL)  -- The minimum concentration of an
           analyte that can be identified, measured and reported with 99% confidence that the
           analyte concentration is greater than zero.

    3.13   MATERIAL SAFETY DATA (MSDS) -- Written information provided by vendors
           concerning a chemical's toxicity, health hazards, physical properties, fire, and
           reactivity data including storage, spill,  and handling precautions.

    3.14   CONTINUING CALIBRATION CHECK (CCC) - A calibration standard
           containing one or more method analytes,  which is analyzed periodically to verify the
           accuracy of the existing calibration for those analytes.

    3.15   MINIMUM REPORTING LEVEL (MRL) - The minimum concentration of an
           analyte that should be reported.  This concentration is determined by the background
           level of the analyte in the LRBs and the sensitivity of the method to the analyte. The
           MRL will be at or near the concentration of the lowest  calibration standard.

4.   INTERFERENCES

    4.1    Method interferences may be caused by contaminants in laboratory air, solvents,
           reagents (including reagent water), glassware, sample bottles and caps, and other
           sample processing hardware that lead to discrete artifacts and/or elevated baselines
           in the chromatograms.  All items such as these must be routinely  demonstrated to be
           free from interferences (less than 1/2 the MRL) under the conditions of the analysis
           by analyzing laboratory reagent blanks as described in Section 9.3. Subtracting
           blank values from sample results is not permitted.

           4.1.1     Before attempting analyses by this method, the analyst  must obtain a
                    source of sufficiently pure reagent water.  This water must not contain
                    target carbonyl compounds or interferences at a concentration greater than
                    1/2 of the MRL. The most likely interferents are formaldehyde and
                    acetaldehyde in the reagent water.  The most successful techniques for
                    generating aldehyde free water are exposure to UV light, or distillation
                    from permanganate.

-------
       4.1.2     Commercially available systems for generating reagent grade water have
                proved adequate, if a step involving exposure to UV light is included. For
                the data presented in this method, a Millipore Elix 3 reverse osmosis
                system followed by a Milli-Q TOC Plus polishing unit provided reagent
                water with background levels of 1 |ig/L or less for each method analyte.
                Other researchers have reported typical blank values of 1-3 |ig/L.(4'5)

       4.1.3     Distillation of reagent water from acidified potassium permanganate has
                been reported as an effective method of eliminating background levels of
                aldehydes.(3) Distill  500 mL of reagent water to which 64 mg potassium
                permanganate and 1  mL concentrated sulfuric acid have been added. In
                our laboratory, this procedure reduced formaldehyde levels to
                approximately 3 |ig/L.

       4.1.4     It may be necessary to purchase reagent grade water. If acceptably clean
                reagent grade water is purchased, care must also be taken to protect it from
                contamination caused by contact with laboratory air.

4.2    Formaldehyde is typically present in laboratory  air and smaller amounts of other
       aldehydes may also be found.  Care should be taken to minimize exposure of
       reagents and sample water with laboratory air.  Because latex is a potential aldehyde
       contaminant source, protective gloves should not contain latex. Nitrile gloves, such
       as N-Dex Plus, are acceptable.  Bottle caps should be made of polypropylene.
       Commonly used phenolic resin caps must be avoided because they can introduce
       formaldehyde contamination into samples.

4.3    Reagents must also be free from contamination.  Many brands of solvents may
       contain trace amounts of carbonyl compounds.

4.4    Glassware must be scrupulously cleaned by detergent washing with hot water, and
       rinses with tap water and distilled water.  Glassware should then be drained, dried,
       and heated in a laboratory oven at 130 °C for several hours before use. Solvent
       rinses with methanol or acetonitrile, followed by air drying,  may be substituted for
       the oven heating.  After cleaning, glassware should be stored in a clean environment
       to prevent any  accumulation of dust or other contaminants.

4.5    Matrix interferences may be caused by contaminants that are co-extracted  from the
       sample.  The extent of matrix interferences will  vary considerably from source to
       source, depending upon the nature and diversity of the matrix being sampled.

4.6    An interferant that elutes just prior to the acetaldehyde (E) isomer peak on the
       primary column is typically observed in chlorinated or chloraminated waters. If this
       peak interferes with the integration of the acetaldehyde  (E) isomer peak, then
       acetaldehyde should be quantified using only the acetaldehyde (Z) isomer, or from
       the confirmation column data.

-------
5.   SAFETY

    5.1    The toxicity or carcinogen!city of each reagent used in this method has not been
           precisely defined; however, each chemical compound should be treated as a potential
           health hazard.  From this viewpoint, exposure to these chemicals must be reduced to
           the lowest possible level by whatever means available.  The laboratory is responsible
           for maintaining a current awareness file of OSHA regulations regarding the safe
           handling of the chemicals specified in this method.  A reference file of material
           safety data sheets should also be made available to all personnel involved in the
           chemical analysis. Additional references to laboratory safety are available.(6"9)

    5.2    Formaldehyde and acetaldehyde have been tentatively classified as known or
           suspected human or mammalian carcinogens. Glyoxal and methyl glyoxal have been
           shown to be mutagenic in in-vitro tests.(3)

    5.3    Although hydrogen can be used as a carrier gas safely, the potential for fire or
           explosion does exist if the gas system is mishandled. If you are unsure of the safety
           guidelines for using hydrogen as a carrier  gas, seek advice from your instrument
           manufacturer regarding its use.

6.   EQUIPMENT AND SUPPLIES (All specifications  are  suggested.  Brand names and/or
    catalog numbers are included for illustration only.)

    6.1    SAMPLE CONTAINERS -- Grab Sample Bottle (aqueous samples) -- 30 mL amber
           glass, screw cap bottles and caps equipped with PTFE-faced silicon septa.  Screw
           caps should be polypropylene Typical phenolic resin caps should be avoided due
           to the possibility of sample contamination from formaldehyde   Prior to use,
           wash bottles and septa according to Section 4.4.

    6.2    VIALS -- 8 mL or 12 mL vials for the acid wash step (Section 11.1.10),  and GC
           autosampler vials, both types must  be glass with PTFE-faced silicon-lined
           polypropylene caps.

    6.3    VOLUMETRIC FLASKS — various sizes used for preparation of standards.

    6.4    BALANCE — Analytical, capable of accurately weighing to the nearest 0.0001 g.

    6.5    WATER BATH or HEATING BLOCK - Capable of maintaining 3 5 ± 2 °C

    6.6    GAS CHROMATOGRAPH - Capillary Gas Chromatograph (Hewlett Packard 6890
           or equivalent), equipped for fast gas chromatography. Modifications will include a
           high pressure (>50 psi) split/splitless injector, fast temperature ramp (30 °C/minute)
           oven and a low volume  (150 (jL) micro ECD detector.  Additionally, a data system
           capable of fast sampling (20 points/peak)  is required.

           6.6.1    Primary Column -  10 m x 0.10 mm J&W DB-5, 0.10 |im film thickness (or
                    equivalent).

-------
           6.6.2    Confirmation Column -- 10 m x 0.10 mm Alltech AT- 1701, 0.10 |im film
                   thickness (or equivalent)

           6.6.3    Straight injection port liners (< 2 mm i.d.) packed with a central 2 cm plug
                   of silanized glass wool. Hewlett Packard PN 5181-3317 or equivalent.

7.   REAGENTS AND STANDARDS

    7.1    Reagent grade or better chemicals should be used in all tests. Unless otherwise
           indicated, it is intended that all reagents shall conform to the specifications of the
           Committee on Analytical Reagents of the American Chemical Society, where such
           specifications are available.  Other grades may be used, provided it is first
           ascertained that the reagent is of sufficiently high purity to permit its use without
           lessening the accuracy of the determination.

    7.2    REAGENT WATER — Reagent water as free as possible from interferences and
           contamination is critical to the success of this method.  See Section 4.1.

    7.3    ACETONITRILE - High purity, demonstrated to be free of analytes and
           interferences.

    7.4    HEXANE — High purity, demonstrated to be free of analytes and interferences:
           B&J Brand, GC2 grade or equivalent.

    7.5    POTASSIUM HYDROGEN PHTHALATE (KHP) - ACS Grade or better.

    7.6    0-(2,3,5,6-PENTAFLUOROBENZYL)-HYDROXYLAMINE HYDROCHLORIDE
           (PFBHA) - 98+%, Aldrich cat. # 19,448-4. (Store in a desiccator - Do not
           refrigerate).

    7.7    SULFURIC ACID - ACS Grade or better.

    7.8    COPPER SULFATE PENTAHYDRATE - ACS Grade or better.

    7.9    AMMONIUM CHLORIDE, NH4C1 or AMMONIUM SULF ATE, (NH4)2SO4 - ACS
           grade or better.

    7.10   SOLUTIONS

           7.10.1   PFBHA REAGENT — Prepare a fresh 15 mg/mL solution in reagent water
                   daily. Prepare an amount appropriate to the number of samples to be
                   derivatized.  One mL of solution is added per sample.  For example,  if 14
                   sample vials are being extracted, prepare 15 mL of solution.  For a 15 mL
                   volume of solution, weigh  0.225 grams of PFBHA into a dry 40 mL vial,
                   add 15 mL water and shake to dissolve.

-------
       7.10.2   0.2 N SULFURIC ACID - Add 5 mL of concentrated sulfuric acid to 900
               mL of reagent water.

7.11    STOCK STANDARD SOLUTIONS - When a compound purity is assayed to be
       96% or greater, the weight can be used without correction to calculate the
       concentration of the stock standard.

       7.11.1   INTERNAL STANDARD (IS) - 1,2-DIBROMOPROPANE, 98+%
               purity. An alternate compound may be used as the IS at the discretion of
               the analyst. If an alternate is selected, an appropriate concentration will
               need to be determined.

               7.11.1.1 INTERNAL STANDARD STOCK SOLUTION, (10,000 |ig/mL)
                       — Accurately weigh approximately 0.1 gram to the nearest
                       O.OOOlg, into a tared 10 mL volumetric flask containing hexane
                       up to the neck. After determining weight difference, fill to mark
                       with hexane. Stock solutions can be used for up to 6 months
                       when stored at -10 °C.

               7.11.1.2 INTERNAL STANDARD FORTIFIED EXTRACTION
                       SOLVENT, 400 |ig/L in hexane — This is the solvent used to
                       extract the derivatized samples. The internal standard is added to
                       the solvent prior to performing the extraction.  The volume of this
                       solvent to be prepared should be determined by the sample
                       workload.  The following example illustrates preparation of 1 L
                       of fortified solvent. If fewer samples are to be analyzed each
                       month, prepare smaller batches of working solvent.  Add 40 jiL of
                       internal standard stock solution directly to 1 L of hexane in a
                       volumetric flask. Cap flask and invert three times to ensure
                       thorough mixing. Transfer to 1 L storage bottle with PTFE-faced
                       silicon lined cap. This solution can be used up to 4 weeks. As a
                       check, run a sample of this working solvent on the FGC before
                       the first extraction of aqueous samples. Have enough working
                       solvent available to extract all calibration and aqueous samples in
                       each extraction set.  Never use two different batches of working
                       solvent for one set of extractions.

       7.11.2   SURROGATE (SUR) -- 2',4',5' -TRIFLUOROACETOPHENONE
               This compound was found to be an appropriate surrogate analyte for these
               analyses. However, the chromatograms for this analysis are very crowded,
               and all possible matrix interferences  cannot be anticipated.  An alternate
               carbonyl compound may be selected  as the surrogate analyte if matrix
               interferences or chromatographic problems are encountered.  Any
               surrogate analyte selected must form an oxime derivative, because one of
               its purposes is to monitor the derivatization process.  If an alternate
               surrogate is selected, its concentration may also be adjusted to meet the
               needs of the laboratory.

-------
         7.11.2.1 SURROGATE STOCK SOLUTION, 10,000 |ig/mL - Accurately
                 weigh approximately 0.1 gram SUR to the nearest O.OOOlg, into a
                 10 mL tared volumetric flask containing room temperature (25 °C)
                 acetonitrile up to the neck. After determining weight difference,
                 fill to mark with acetonitrile. Stock solutions can be used for up
                 to 6 months when stored at -10 °C or less.

         7.11.2.2 SURROGATE ADDITIVE SOLUTION, 20 |ig/mL -- Dilute the
                 surrogate stock solution to 20 |ig/mL in acetonitrile.  This
                 solution can be used up to 3 months when stored at 4 °C or less.

7.11.3    STOCK STANDARD SOLUTION (SSS)
         Prepare stock standard solutions for each analyte of interest at a
         concentration of 1 to 10 mg/mL.  Acetonitrile should be used as the
         solvent for all analytes except glyoxal.  Glyoxal standards should be
         prepared using a volumetric 90:10 acetonitrile:water mixture  due to the
         limited solubility of glyoxal in pure acetonitrile. Method analytes may be
         obtained as neat materials or as ampulized solutions from commercial
         suppliers.  The stock standard solutions should be stored at -10 °C or less
         and protected from light.  Standards prepared in this manner were stable
         for at least 60 days.  Laboratories should use standard QC practices to
         determine when their standards need to be replaced.

         7.11.3.1 For analytes which are solids in their pure form, prepare stock
                 standard solutions by accurately weighing approximately 0.1
                 gram of pure material to the nearest O.OOOlg in a 10 mL
                 volumetric flask. Dilute to volume with solvent.

         7.11.3.2 Stock standard solutions for analytes which are liquid in their
                 pure form at room temperature can be accurately prepared in the
                 following manner.

                     7.11.3.2.1    Place about 9.8 mL of solvent into a 10-mL
                                  volumetric flask. Allow the flask to stand,
                                  unstoppered, for about 10 min to allow solvent
                                  film to evaporate from the inner walls of the
                                  volumetric, and weigh to the nearest 0.0001
                                  gram.

                     7.11.3.2.2    Use a 100-jiL syringe and immediately add
                                   100 jiL of standard material to the flask by
                                  keeping the syringe needle just above the
                                  surface of the solvent.  Be sure the standard
                                  material falls dropwise directly into the
                                  solvent without contacting the inner wall of
                                  the volumetric.
                            10

-------
                     7.11.3.2.3     Reweigh, dilute to volume, stopper, then mix
                                  by inverting several times. Calculate the
                                  concentration in milligrams per milliliter from
                                  the net gain in weight.

7.11.4    PRIMARY DILUTION STANDARD (PDS) - The PDS for this method
         should include all method analytes of interest. The PDS is prepared by
         combining and diluting stock standard solutions with acetonitrile to a
         concentration of 100 |ig/mL.  Store at -10 °C or less and protect from light.
         Standards prepared in this manner were stable for at least 60 days.
         Laboratories should use standard QC practices to determine when their
         standards need to be replaced. This primary dilution standard is used to
         prepare calibration spiking solutions, which are prepared at 5
         concentration levels for each analyte, and are used to spike reagent water
         to prepare the aqueous calibration standards.

7.11.5    CALIBRATION SPIKING SOLUTIONS -- Five calibration spiking
         solutions are prepared, each at a different concentration, and are used to
         spike reagent water to prepare the calibration standards. The calibration
         spiking solutions are prepared from the PDS.  Store the calibration  spiking
         solutions at -10 °C or less and protect from light. Solutions prepared in
         this manner were stable for at least 60 days. Laboratories should use
         standard QC practices to determine when solutions need to be replaced.
         An example of how the calibration spiking solutions are prepared is given
         in the following table.  Modifications of this preparation scheme may be
         made to meet the needs of the laboratory.

         PREPARATION OF CALIBRATION SPIKING SOLUTIONS

Calibration
Level


1
2
3
4
5

PDS
Concentration,
(|ig/mL)

100
100
100
100
100

Volume PDS
Standard,
(jiL)

250
500
1000
1500
2000
Final Volume
Calibration
Spike
Solution, (mL)

5
5
5
5
5
Final
Concentration
Calibration
Spike Solution,
(|ig/mL)
5
10
20
30
40
7.11.6    PROCEDURAL CALIBRATION STANDARDS -- A designated amount
         of each calibration spiking solution is spiked into five separate 20 mL
         aliquots of reagent water in a 30 mL sample container, to produce
         aqueous calibration standards. The reagent water used to make the
         calibration standards should contain the preservation reagents described in
                            11

-------
                    Section 8.1.2 (ammonium chloride or ammonium sulfate at 500 mg/L and
                    copper sulfate pentahydrate at 500 mg/L). Aqueous calibration standards
                    are processed and analyzed according to the procedures in Section 11.
                    Resulting data are used to generate a calibration curve. An example of the
                    preparation of aqueous calibration standards is given below. The lowest
                    concentration calibration standard must be at or below the MRL.
                    Modifications of this preparation scheme may be made to meet the needs
                    of the laboratory. Preparing aqueous calibration standards using varying
                    volumes of one calibration spiking solution is an acceptable alternative to
                    the example below.

                    PREPARATION OF PROCEDURAL (AQUEOUS)
                    CALIBRATION STANDARDS

Calibration
Level


1
2
3
4
5
Calibration Spike
Solution
Concentration,
(|ig/mL)

5
10
20
30
40
Volume
Calibration
Spike
Solution,
(ML)
20
20
20
20
20
Final Volume
Calibration
Standard
(mL)

20
20
20
20
20
Final
Concentration
Calibration
Standard
(W?/L)
5
10
20
30
40
8.   SAMPLE COLLECTION. PRESERVATION. AND STORAGE

    8.1    SAMPLE VIAL PREPARATION

           8.1.1     Grab samples must be collected in accordance with conventional sampling
                    practices (7) using amber glass 30 mL containers with PTFE-lined screw-
                    caps, or caps with PTFE-faced silicon septa.

           8.1.2     Prior to shipment to the field, 15 mg of copper sulfate pentahydrate must
                    be added to each bottle. This material acts as a biocide to inhibit
                    bacteriological decay of method analytes. If samples to be collected
                    contain free chlorine, then 15 mg of ammonium chloride or ammonium
                    sulfate  must also be added to the bottle prior to sample collection. The
                    ammonium compound will react with the free chlorine to form
                    monochloramine, and retard the formation of additional carbonyl
                    compounds. Add these materials as dry solids to the sample bottle. The
                    stability of these materials in concentrated aqueous solution has not been
                    verified.

                    NOTE: Aldehydes have been demonstrated to be extremely susceptible to
                            microbiological decay. The use of other chlorine reducing agents
                                       12

-------
                            such as sodium thiosulfate or ascorbic acid, has also been shown
                            to produce invalid data. Proper sample collection and
                            preservation is important to obtaining valid data. The data in
                            Section 17, Table 7 illustrates the importance of proper sample
                            preservation.

8.2  SAMPLE COLLECTION

           8.2.1     Fill sample bottles to just overflowing but take care not to flush out the
                    sample preservation reagents.  The capped sample should be head-space
                    free.

           8.2.2     When sampling from a water tap, remove the aerator so that no air bubbles
                    will be trapped in the sample. Open the tap,  and allow the system to flush
                    until the water temperature has stabilized (usually about 3-5 minutes).
                    Collect samples from the flowing system.

           8.2.3     When sampling from an open body of water, fill a 1 quart wide-mouth
                    bottle or 1L beaker with sample from a representative area,  and carefully
                    fill sample bottles from the container.

           8.2.4     After collecting the sample, cap carefully to avoid spillage,  and agitate by
                    hand for 1 minute.

8.3  SAMPLE STORAGE/HOLDING TIMES

           8.3.1     Samples must be iced or refrigerated at 4 + 2 °C and maintained at these
                    conditions away from light until extraction.  Samples must be extracted
                    within 7 days of sampling.  However, since aldehydes are subject to decay
                    in stored samples, all samples should be derivatized and extracted as soon
                    as possible.

                    NOTE: A white or blue precipitate is likely to occur.  This is normal and
                            does not indicate any problem with sample collection or storage.

           8.3.2     Extracts (Section 11.1.11) must be stored at 4  + 2 °C away from light in
                    glass vials with PTFE-faced silicon-lined caps. Extracts must be analyzed
                    within 14  days of extraction.

     8.4    FIELD REAGENT BLANKS -  Processing of a field reagent blank (FRB) is
           required along with  each sample set.  A sample set is composed of the samples
           collected from the same general sampling site at approximately the same time. Field
           reagent blanks are prepared at the laboratory before sample vials are sent to the field.
           At the laboratory, fill a sample container with reagent water (Section 7.2), add
           sample preservatives as described in Section 8.1.2, seal and ship to the sampling site
           along with the empty sample containers.  FRBs should be confirmed to be free (less
           than 1/2 the MRL) of all method analytes prior to shipping them to the field.  Return

                                       13

-------
           the FRB to the laboratory with filled sample bottles. DO NOT OPEN THE FRB
           AT THE  SAMPLING SITE. If any of the analytes are detected at concentrations
           equal to or greater than 1/2 the MRL , then all data for the problem analyte(s) should
           be considered invalid for all  samples in the shipping batch.

9.    QUALITY CONTROL

     9.1    Each laboratory that uses this method is required to operate a formal quality control
           (QC) program. Minimum QC requirements are initial demonstration of laboratory
           capability (which includes calculation of the MDL), analysis of laboratory reagent
           blanks, laboratory fortified blanks, field reagent blanks, laboratory fortified sample
           matrices, and QC samples. Additional QC practices are encouraged.

     9.2    INITIAL DEMONSTRATION OF CAPABILITY (IDC) -- Requirements for the
           initial demonstration of laboratory capability are described in the following sections
           and summarized in  Section 17, Table 8.

           9.2.1     Initial demonstration of low system background. (Section 9. 3)

           9.2.2     Initial demonstration of precision.  Prepare, derivatize, extract, and analyze
                     4-7 replicate LFBs fortified at 20 ug/L, or other mid-range concentration,
                     over a period of at least 2 days.  Generating the data over a longer period
                     of time, e.g., 4 or 5 days may produce a more realistic indication of day to
                     day  laboratory performance.  The relative standard deviation (RSD) of the
                     results of the replicate analyses must be less than 20%.

           9.2.3     Initial demonstration of accuracy. Using the same set of replicate data
                     generated for Section 9.2.2, calculate average recovery. The average
                     recovery of the replicate values must be within ± 20% of the true value.

           9.2.4     MDL determination(1'2). Replicate analyses for this procedure should be
                     done over at least 3 days (both the sample derivatization/extraction and the
                     GC  analyses should be done over at least 3 days). Prepare at least 7
                     replicate LFBs at a concentration estimated to be near the MDL. This
                     concentration may be estimated by selecting a concentration at 2-5X the
                     noise level. Analyze the seven replicates through all steps of Section 11.
                     Calculate the MDL
                            MDL=St(n.1 1. alpha = 0.99)

                    where:

                            Vu-aipha = 0.99)= Student's t value for the 99% confidence level
                            with n-1 degrees of freedom
                            n = number of replicates
                            S = standard deviation of replicate analyses.

                                        14

-------
                NOTE: Do not subtract blank values when performing MDL calculations.

       9.2.5     Minimum Reporting Level (MRL) — Although an MDL can be calculated
                for analytes that commonly occur as background contaminants, the
                calculated MDLs should not be used as the MRL for each analyte. Method
                analytes that are seen in the background (typically formaldehyde,
                acetaldehyde) should be reported as present in field samples, only after
                careful evaluation of the background levels. It is recommended that a
                MRL be established at the mean LRB concentration + 3o, or three times
                the mean LRB concentration, whichever is greater.  This value should be
                calculated over a  period of time, to reflect variability in the blank
                measurements. It is recommended that this value be used as a minimum
                reporting level in order to avoid reporting false positive results.

9.3    LABORATORY REAGENTS BLANKS (LRB) -- Each time a set of samples is
       extracted or reagents are changed, a LRB must be analyzed.  If within the retention
       time window of any analyte, the LRB produces a peak that would prevent the
       determination of that analyte, determine the source of contamination and eliminate
       the interference before processing samples.  Because background contamination is a
       significant problem for several method analytes, it is highly recommended that the
       analyst maintain a historical record of LRB data. If target analytes are detected in
       the LRB at concentrations equal to  or greater than 1/2 the MRL (Section 9.2.5), then
       all data for the problem analyte(s) should be considered invalid for all samples in the
       extraction batch.

9.4    CONTINUING CALIBRATION CHECK/LABORATORY FORTIFIED BLANK --
       Since this methodology is based on procedural standard calibration, a LFB and the
       calibration check sample (CCC) are prepared and analyzed in the same manner.
       Laboratory fortified blank QC requirements are therefore omitted. Calibration
       procedure options and the QC acceptance criteria associated with them are fully
       described in Section 10.3. Please refer to that section for these criteria.

9.5    INTERNAL STANDARD—The analyst must monitor the IS  response peak area of
       all injections during each analysis day. A mean IS response is determined from the
       five point calibration curve. The IS response for any chromatographic run should not
       deviate from this mean IS response by more than 30%. If a deviation greater than
       30% occurs with an individual extract, inject a second aliquot of that extract.

       9.5.1     If the reinjected aliquot produces an  acceptable internal standard response,
                report results for that aliquot.

       9.5.2     If a deviation of greater than 30% is  obtained for the reinjected extract, the
                analyst should check the calibration by analyzing the most recently
                acceptable calibration standard. If the calibration standard fails the criteria
                of Section 10.3, recalibration is in order per Section 10. If the calibration
                standard is acceptable, extraction of the sample should be repeated

                                   15

-------
                provided the sample is still within the holding time. Otherwise, report
                results obtained from the reinjected extract, but annotate as suspect.

9.6    SURROGATE RECOVERY-The surrogate standard is fortified into the aqueous
       portion of all calibration standards, samples, FRBs and LRBs. The surrogate is a
       means of assessing method performance from derivatization to final
       chromatographic measurement.

       9.6.1     When surrogate recovery from a sample, blank, or CCC is <70% or
                >130%, check (1) calculations to locate possible errors, (2) standard
                solutions for degradation, (3) contamination,  and (4) instrument
                performance. If those steps do not reveal the cause of the problem,
                reanalyze the extract.

       9.6.2     If the extract reanalysis meets the surrogate recovery criterion, report only
                data for the reanalyzed extract.

       9.6.3     If the extract reanalysis fails the 70-130% recovery criterion, the analyst
                should check the calibration by analyzing the most recently acceptable
                calibration standard. If the calibration standard fails the criteria of Section
                9.6.1, recalibration is in  order per Section 10. If the calibration standard is
                acceptable, it may be necessary to extract another aliquot of sample if
                sample holding time has not been exceeded. If the sample re-extract also
                fails the recovery criterion, report all data for that sample as suspect.

9.7    LABORATORY FORTIFIED SAMPLE MATRIX (LFM)

       9.7.1     Within each analysis set, a minimum of one field sample is fortified as a
                LFM for every 20 samples analyzed. The LFM is prepared by spiking a
                sample with an appropriate  amount of the calibration standard. The
                concentrations 5, 10, and 20 |ig/L are suggested spiking concentrations.
                Select the spiking concentration that is closest to, but greater than the
                concentration in the unfortified sample. Use  historical data or rotate
                through the designated concentrations to select a fortifying concentration.
                Selecting a duplicate vial of a sample that has already been analyzed, aids
                in the selection of appropriate spiking levels.

       9.7.2     Calculate the percent recovery (R) for each analyte, after correcting the
                measured concentration, A,  from the fortified sample for the background
                concentration, B, measured  in the unfortified sample, i.e.,
                                      c

                where C is the fortified concentration. Compare these values to control limits
                appropriate for reagent water data collected in the same fashion.
                                    16

-------
           9.7.3     Recoveries may exhibit a matrix dependence. For samples fortified at or
                    above their native concentration, recoveries should range between 70 -
                    130%. If the accuracy of any analyte falls outside the designated range,
                    and the laboratory performance for that analyte is shown to be in control,
                    the accuracy problem encountered with the fortified sample is judged to be
                    matrix related, not system related.  The result for that analyte in the
                    unfortified sample is labeled suspect/matrix to inform the data user that the
                    results are suspect due to matrix effects. Repeated failure to meet the
                    suggested recovery criteria indicates potential problems with the extraction
                    procedure and should be investigated.

    9.8    FIELD DUPLICATES — Within each analysis batch, a minimum of one field sample
           should be analyzed in duplicate. Duplicate sample analyses serve as a check on
           sampling and laboratory precision.

           9.8.1     Calculate the relative percent difference (RPD) for duplicate
                    measurements (Ldl and Ld2)  as shown below.
                                RPD=     d~d     *10Q
           9.8.2    Relative percent differences for laboratory duplicates should fall in the
                    range of ± 30 %. Greater variability may be observed for target analytes
                    with concentrations near their MRL.

    9.9    QUALITY CONTROL SAMPLE (QCS) -- At least quarterly, analyze a QCS from
           an external source. If measured analyte concentrations are not of acceptable
           accuracy (70-130% of the expected value), check the entire analytical procedure to
           locate and correct the problem source.

    9.10   ASSESSING (Z/E) RATIOS -- In addition to monitoring analyte response from
           CCC/LFB, the ratio of the peak areas of each isomer pair should be monitored.
           When samples and standards are processed and analyzed by exactly the same
           procedure, the ratio of the (Z/E) isomers produced by each method analyte will be
           reproducible. This information can be used as a QC check to avoid biased results
           caused by an interferant with one isomer of the pair. Calculate and record the ratio of
           the peak area of the first eluting isomer (designated (E)) to the second eluting isomer
           (designated (Z)). This ratio will be used in data evaluation Section 12.4.

10. CALIBRATION AND STANDARDIZATION

    10.1   Demonstration and documentation of acceptable  initial calibration is required before
           any samples are analyzed, and is required intermittently throughout sample analysis.
           After initial calibration is successful, the analyst may choose one of two options for
           maintaining on-going calibration. The first option is to verify the initial calibration

                                        17

-------
       daily using a minimum of 2 calibration standards. The other option is daily
       calibration of the method with all 5 calibration standards. These options are further
       described in Section 10.3.
10.2   INITIAL CALIBRATION CURVE

       10.2.1   Establish FGC operating parameters equivalent to the suggested
               specifications in Section 17, Table 1. The GC system must be calibrated
               using the internal standard (IS) technique. Other GC columns or GC
               conditions may be used if equivalent or better performance can be
               demonstrated (Section 1.3).

       10.2.2   Five calibration standards are recommended to calibrate over the range of
               approximately 5-40 |ig/L.  The lowest level standard will depend upon the
               level of blank contamination for each analyte (Section 7.11.6).

       10.2.3   Prepare each calibration standard by the procedural standard calibration
               method.  Method analytes are fortified into reagent water and carried
               through the entire extraction and derivatization procedure described in
               Section 11.

       10.2.4   Inject 1 (jL of each calibration standard extract into the FGC and tabulate
               peak area response and concentration for each analyte and the internal
               standard. NOTE: The formaldehyde peak will be much larger (for the
               same concentration) than the other analyte peaks.  The formaldehyde peak
               may need to be attenuated on  some instruments/data systems to avoid
               signal saturation.

       10.2.5   (Z/E) ISOMERS — Two isomers, referred to as (E) and (Z), are formed for
               most asymmetrical carbonyl compounds derivatized with PFBHA.
               Chromatographic resolution is usually obtained with the columns
               suggested in  Section 6.6 for acetaldehyde, propanal, butanal, pentanal,
               hexanal, heptanal, and octanal, (see chromatograms in Section  17, Figure 1
               and Figure 2). With dicarbonyl species such as glyoxal and methyl
               glyoxal,  (E) and (Z) isomerism occurs from oxime formation with both
               carbonyl groups, increasing the number of isomers.  The demonstration
               data included in this method use two distinct isomer peaks each for
               glyoxal and methyl  glyoxal.  Use one of the following methods for both
               calibration and quantitation of each method analyte.

               (a)  Use the sum of the isomer peak areas  for each constituent for both
                    calibration and quantitation.

               (b)  Use the peak area of each individual isomer to independently
                    calculate a concentration for each isomer. Then average the amount
                    of the two isomers to report one value for the analyte.

                                   18

-------
        10.2.6   Generate a calibration curve for each analyte by plotting the area ratios
                (Aa/AjS) against the concentration ratios (Ca/Cis) of the five calibration
                standards where:

                     Aa  is the peak area of the analyte (or analyte isomer pair),
                     A;s  is the peak area of the internal standard,
                     Ca  is the concentration of the analyte, and
                     Cis  is the concentration of the internal standard.

        10.2.7   This curve must always be forced through zero and can be defined as
                either first or second order.  Forcing zero allows for a better estimate of the
                background level of method analytes.

        10.2.8   A data system is required to collect the chromatographic data, to calculate
                relative response factors, and calculate either linear or second order
                calibration curves.

        10.2.9   VERIFICATION OF CALIBRATION STANDARD MATERIALS --
                Analyze a LFB prepared from standard materials from a source other than
                those used to prepare the initial calibration curve (Sections 3.8, and 9.9).
                Calculate the concentration of this QCS from the calibration curve. The
                calculated concentration of the QCS must agree within 70-130% of its true
                value.  This step verifies the validity of calibration standard materials and
                the calibration curve prior to sample analyses.

10.3     OPTIONS FOR ON-GOING CALIBRATION

        The time, temperature, pH, and PFBHA concentration will all affect the rate,
        efficiency and reproducibility of the derivatization reaction. It is critical that those
        parameters be controlled. Calibration frequency will depend upon the laboratory's
        ability to control these parameters so that continuing calibration check standard
        criteria can be met. Some laboratories may find it more productive to prepare and
        analyze a calibration curve with each batch of samples. A batch of samples for this
        methodology should not exceed 20 samples, including field samples, FRBs,
        laboratory duplicates, and fortified sample matrices.

        10.3.1   CONTINUING CALIBRATION CHECK (CCC) OPTION-The analyst
                must periodically verify calibration during the analysis of samples in order
                to ensure accuracy of analytical results.  Prepare a minimum of one low-
                level (suggested concentration 2-5 ug/L) and one mid-level (suggested
                concentration 10-30 |ig/L) calibration standard with each batch of samples.
                Verify calibration using these two  standards, prior to analyzing any of the
                sample extracts from the batch. In addition, reanalyze one of these two
                standard extracts after every tenth sample extract, and after the last sample
                in an analysis batch to ensure instrument stability throughout the analysis
                batch.  Recovery must be within 70-130% of the true value for the mid-

                                   19

-------
                    level standard, and within 50-150% of the true value for the low-level
                    standard.

            10.3.2   DAILY CALIBRATION OPTION - The analyst may choose to create a
                    new calibration curve for each batch of samples by preparing and
                    analyzing a standard at all five calibration concentrations, with each batch
                    of samples. If this option is selected, the calibration standard extracts
                    should be analyzed prior to the analysis of sample extracts.  To ensure that
                    sensitivity and performance of the method has not changed significantly
                    between sample batches, or changed since the IDC, the following
                    performance check is required.  The response (peak area) of the internal
                    standard, surrogate and each method analyte in the mid-level standard
                    (suggested concentration 10-30 |ig/L), must be within 50-150% of the
                    mean peak area for that analyte in the initial demonstration of precision
                    replicates (Section 9.2.2).  One of the calibration standard extracts must be
                    reanalyzed after every tenth sample extract,  and after the last sample in an
                    analysis batch to ensure instrument stability throughout the analysis batch.
                    Recovery must be within 70 to 130% of the true value for mid- and high-
                    level calibration standards, and within 50-150% of the true value for the
                    low-level standard (suggested concentration 2-5 |ig/L).

11  PROCEDURE

    11.1    SAMPLE EXTRACTION — Once samples have been opened, process the samples
            straight through to step 11.1.11. There is no known "safe" stopping point once
            sample processing has begun.  Samples are derivatized and extracted in the sample
            bottle in which they were collected.  Transferring the sample to another container
            for derivatization and extraction has been shown to cause a loss of method analytes.

            11.1.1   Remove the samples from storage and allow them to equilibrate to room
                    temperature.

            11.1.2   Remove 10 mL of sample and discard. Mark the level of the remaining
                    sample volume on the outside of the bottle, for later sample volume
                    determination.

            11.1.3   Add 200 mg KHP to adjust the sample pH to approximately 4.

            11.1.4   Add 20 |iL surrogate solution (Sect 7.11.2.2).

            11.1.5   Add 1 mL of freshly prepared PFBHA Reagent as per Section 7.10.1.  Cap
                    and swirl gently to mix.

            11.1.6   Place all samples in a constant-temperature water bath set at 35 ± 2 °C for
                    2 hours. Remove vials and cool to room temperature for 10 minutes.
                                       20

-------
       11.1.7   To each vial add approximately 0.05 mL (2 to 4 drops) of concentrated
                sulfuric acid. This prevents the extraction of excess reagent, which will
                cause chromatographic interferences.

       11.1.8   Add 4 mL of hexane that contains the internal standard (Section 7.11.1.2).

       11.1.9   Shake manually for 3 minutes.  Let stand for approximately 5 minutes to
                permit phases to separate.

       11.1.10  Draw off hexane layer (top layer) using a  clean disposable Pasteur pipette
                for each sample into a smaller 8 mL vial containing 3 mL 0.2 N sulfuric
                acid.  Shake for 30 seconds and let stand for 5 minutes for phase
                separation. NOTE: This acid wash step further reduces the reagent and
                other interferants from the final extract.

       11.1.11  Draw off top hexane layer using another clean, disposable pipette for each
                sample and place in two 1.8 mL autosampler vials per sample. Store extra
                autosampler vials as a backup extract. Extracts may be stored for up to 14
                days at 4 °C.

       11.1.12  Sample Volume Determination — Discard  remaining water sample and
                hexane in each sample bottle. Fill with water to the level indicated by the
                mark made in Section 11.1.2. Pour the water into a 25 mL graduated
                cylinder and measure the volume to the nearest mL.  Record the sample
                volume for each sample.

                Alternately, if a laboratory has control over the brand and style of the 30
                mL sample bottles being used, the exact volume of a number of bottles
                from the same manufacturer and lot may be measured, and the average
                bottle volume minus 10 mL may be used as the sample volume for all
                samples using the same lot of sample bottles. A minimum of 10 % of the
                sample bottles obtained from the same manufacturer,  from the same lot
                should be measured.

11.2    FAST GAS CHROMATOGRAPHY- This method uses fast gas chromatography
       (FGC) for the analysis of the hexane extracts.  Several important changes from
       "conventional GC" must be made to allow for the rapid analysis of the analytes.
       First the instrument must be capable of providing a fast temperature ramp
       (30°C/minute) oven, a high pressure (>50 psi) split/splitless injector, and a low
       volume (150 (jL ) micro BCD.  Second, the column diameter, length and film
       thickness must all be decreased. Third, the carrier gas must be changed to a highly
       diffusive
       or "fast"  gas such as hydrogen.  Although hydrogen can be used safely as a
       carrier gas, the potential for fire or explosion does exist if the gas system is
       mishandled. If you are unsure of the safety guidelines for using hydrogen as a
       carrier gas, seek advice from your instrument manufacturer regarding its use.
       Finally, strict attention must be paid to established column installation guidelines

                                  21

-------
            with regard to the proper cutting and placement of the capillary columns within the
            instrument.

            In addition to decreasing the column dimensions and changing the carrier gas,
            successful FGC depends on the minimization of extracolumn variance.
            Extracolumn variance refers to any source of bandbroadening other than those
            which occur within the chromatographic column itself. The major source of
            extracolumn variance in a properly designed FGC chromatographic system is the
            injector/injection process. The best chromatographic results  for the oxime
            derivatives have been achieved using glass wool packed small i.d. liners (Section
            6.6.3) combined with low split ratios. Although excellent precision and accuracy
            have been demonstrated using the listed chromatographic conditions (Section 17,
            Table 1.), it possible that the optimum conditions for a specific instrument will need
            to be empirically determined by the user.

            11.2.1   Analyze the extracts by FGC/ECD.  Tables 1 and 2 (Section 17)
                    summarize recommended FGC operating conditions and retention times
                    observed using this method. Figure 1 illustrates the performance of the
                    recommended primary column with the method analytes. Figure 2
                    illustrates the performance of the recommended confirmation column with
                    the method analytes.  Other  GC columns or chromatographic conditions
                    may be used if the requirements of Section 9 are met.

            11.2.2   The width of the retention time window used to make identifications
                    should be based on measurements of actual retention time variations of
                    standards over the course of time. Plus or minus three times the standard
                    deviation of the retention time for a compound can be used to calculate a
                    suggested window size; however the experience of the analyst should
                    weigh heavily in the interpretation of chromatograms.

            11.2.3   If an analyte peak area exceeds the range of the calibration curve, the
                    extract may be diluted with the hexane extraction solvent (that contains the
                    internal standard) and reanalyzed. Incorporate the dilution factor into final
                    concentration calculations. The analyst must not extrapolate beyond the
                    calibration range established.

12. DATA ANALYSIS AND  CALCULATIONS

    12.1    Identify the method analytes in the sample chromatogram by comparing the
            retention time of the suspect peak to the retention time of an analyte peak (or isomer
            peaks) in a calibration standard or the laboratory fortified blank.

    12.2    Calculate the analyte concentrations using the first or second order calibration
            curves generated as described in Section 10.

    12.3    For any analytes that are found, adjust the calculated concentration to reflect the true
            sample volume determined in Section 11.1.12.

                                       22

-------
    12.4    Prior to reporting the data, the chromatogram should be reviewed for any incorrect
            peak identification or poor integration.  If a confirmation column has been used, all
            identifications should be verified using the retention time data from that analysis. In
            addition, the (Z/E) isomer ratio should be within 50% of the ratio observed in
            standards.  If the (Z/E)  ratio does not meet these criteria, it is likely that an
            interferent occurred at  the retention time of one of the isomer peaks.  In this case,
            the amount indicated by the lower of the 2 isomer peaks should be reported.  (This
            may require that the  analyst recalculate the analyte amount using individual isomer
            peaks for quantitation.) If one peak of the isomeric pair is missing, the
            identification is not confirmed and should not be reported.

    12.5    Analyte concentrations are reported in |ig/L.
13. METHOD PERFORMANCE

    13.1    Precision and accuracy data are presented in Section 17.  Data are presented for
            three water matrices: reagent water (Table 3), chlorinated "finished" surface water
            (Table 5), chlorinated "finished" ground water (Table 6).

    13.2    DERIVATIZATION PARAMETERS - This method is a procedural standard
            method that will generate accurate and precise results when used as written. The
            time, temperature, pH, and PFBHA concentration will all affect the rate, efficiency
            and reproducibility of the derivatization reaction. It is critical that those parameters
            be controlled. Calibration frequency will depend upon the laboratory's ability to
            control these parameters. Some laboratories may need to prepare and analyze a
            calibration curve with each batch of samples.  Of all the method analytes, glyoxal,
            methyl glyoxal, benzaldehyde, and cyclohexanone are the most difficult to
            derivatize. Poor sensitivity for any of these compounds indicates that there may be
            a problem with the reaction conditions. Measurements of nonanal, decanal, glyoxal
            and methyl glyoxal appear to be less precise than the measurement of other analytes.

    13.3    The importance of low background levels of formaldehyde and acetaldehyde cannot
            be overemphasized. Some laboratories or reagent waters may also contain
            background amounts of other method analytes. Care must be taken to avoid
            reporting false positive results that result from background contamination.

    13.4    The importance of proper sample collection and preservation also cannot be
            overemphasized. Holding time studies in various matrices showed better than 70%
            recovery of all method analytes when samples were collected, preserved, and stored
            according to Section 8, and analyzed within 7 days. There were variations in the
            recovery of analytes from fortified samples from different matrices. Therefore, it is
            strongly recommended that samples be analyzed as soon as possible after collection.
            The data in Section 17, Table 7 illustrate the dramatic difference between a
            preserved and a non-preserved sample. Although this data was presented as Table 6
                                        23

-------
            of Method 556, revision 1.0, it is included to illustrate the importance of proper
            sample preservation.
14.  POLLUTION PREVENTION

    14.1    This method uses a micro-extraction procedure which requires very small quantities
            of organic solvents.

    14.2    For information about pollution prevention that may be applicable to laboratory
            operations, consult "Less is Better: Laboratory Chemical Management for Waste
            Reduction" available from the American Chemical Society's Department of
            Government Relations and Science Policy, 1155 16th Street N.W., Washington,
            D.C., 20036.

15.  WASTE MANAGEMENT

    15.1    The analytical procedures described in this method generate relatively  small
            amounts of waste since only small amounts of reagents and solvents are used.  The
            matrices of concern are finished drinking water or source water. However, the
            Agency requires that laboratory waste management practices be conducted
            consistent with all applicable rules and regulations, and that laboratories protect the
            air, water, and land by minimizing and controlling all releases from fume hoods and
            bench operations. Also, compliance is required with any sewage discharge permits
            and regulations, particularly the hazardous waste identification rules and land
            disposal restrictions.  For further information on waste management, consult "The
            Waste Management Manual for Laboratory Personnel" also available from the
            American Chemical Society at the address in Section  14.2.

16.  REFERENCES

    1.       Glaser, J.A., D.L. Foerst, G.D. McKee, S.A. Quave, and W.L. Budde, "Trace
            Analyses for Wastewaters," Environ. Sci. Technol. 1981, 15, 1426-1435.

    2.       Definition and procedure for the determination of the method detection limit. 40
            CFRPartl36, Appendix B.

    3.       Standard Method Number 6252B, "PFBHA Liquid-Liquid Extraction Gas
            Chromatographic Method," Standard Methods for the Examination of Water and
            Wastewater. pp. 6-77 to 6-83, American Public Health Assoc., Washington, D.C.,
            1995.

    4.       Sclimenti, M.J., S.W. Krasner, W.H. Glaze, and H.S. Weinberg,"Ozone
            Disinfection By-Products:  Optimization of the PFBHA Derivatization Method for
            the Analysis of Aldehydes," In Advances in Water Analysis and Treatment, Proc.
            AWWA Water Quality Technology Conf. 1990, pp 477-501.
                                       24

-------
5.       Glaze, W.H. and H.S. Weinberg, Identification and Occurrence of Ozonation By-
        Products in Drinking Water. American Water Works Assoc. Research Foundation,
        Denver, CO., 1993, pp!9-22.

6.       "OSHA Safety and Health Standards, General Industry," (29CRF1910).
        Occupational Safety and Health Administration, OSHA 2206, (Revised, Jan. 1976).
7.      ASTM Annual Book of Standards, Part II, Volume 11.01, D3370-82, "Standard
       Practice for Sampling Water," American Society for Testing and Materials,
       Philadelphia, PA, 1986.

8.      "Carcinogens-Working with Carcinogens," Publication No. 77-206, Department of
       Health, Education, and Welfare, Public Health Service, Center for Disease Control,
       National Institute of Occupational Safety and Health, Atlanta, Georgia, August
       1977.

9.      "Safety In Academic Chemistry Laboratories," 3rd Edition, American Chemical
       Society Publication, Committee on Chemical Safety, Washington, D.C.,  1979.
                                   25

-------
17.   TABLES. DIAGRAMS. FLOWCHARTS. AND VALIDATION DATA

TABLE 1. CHROMATOGRAPHIC CONDITIONS AND RETENTION
           TIME DATA FOR THE PRIMARY COLUMN (N=8)
Peak Number
(Figure 1.)
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
Analyte
1,2 dibromopropane (IS)
Formaldehyde
E-Acetaldehyde
Z-Acetaldehyde
E-Propanal
Z-Propanal
E-Butanal
Z-Butanal
E-Pentanal
Z-Pentanal
E-Hexanal
Z-Hexanal
Cyclohexanone
E-Heptanal
Z-Heptanal
2,4,5, trifluoroacetophenone (S)
E-Octanal
Z-Octanal
Benzaldehyde
Nonanal
Decanal
Glyoxal (peak 1)
Glyoxal (peak 2)
Methyl Glyoxal (peak 1)
Methyl Glyoxal (peak 2)
Average
Retention
Time (min)
0.673
1.08
1.46
1.50
1.85
1.88
2.28
2.31
2.73
2.76
3.19
3.21
3.54
3.64
3.65
3.97
4.07
4.08
4.19
4.50
4.91
5.23
5.27
5.29
5.41
Standard
Deviation
3.26E-03
1.67E-03
l.OOE-04
7.08E-04
4.51E-04
4.63E-04
5.07E-04
4.86E-04
3.36E-04
3.27E-04
3.18E-04
1.85E-04
4.33E-04
3.47E-04
1.59E-04
3.18E-04
3.25E-04
4.88E-04
2.13E-04
2.97E-04
2.76E-04
5.33E-04
4.91E-04
2.90E-04
2.85E-04
Relative
Standard
Deviation
0.49%
0.15%
0.01%
0.05%
0.02%
0.02%
0.02%
0.02%
0.01%
0.01%
0.01%
0.01%
0.01%
0.01%
0.00%
0.01%
0.01%
0.01%
0.01%
0.01%
0.01%
0.01%
0.01%
0.01%
0.01%
Primary Column:
           DB-5, 10 m x 0.10 mm i.d., 0.10 |j,m film thickness, injector temp. 200 °C, liner 2
           mm with a central 2 cm silanized glass wool plug, injection volume 1 |jL, split ratio
           30:1, constant head pressure @ 32 psi, detector temp. 300 °C, detector make up flow
           20 mL/minute.  Temperature program: 70 °C initial, program at 27 °C/minute to 220
           °C, ballistic heating to 280 °C for burnout and hold at 280 °C for 0.4 minutes.  Data
           collection via HP GC Chemstation at a rate of 50 Hz.

Carrier gas: Hydrogen (UHP)

Detector gas: 95:5 Argon:Methane
                                   26

-------
TABLE 2. CHROMATOGRAPHIC CONDITIONS AND RETENTION
TIME DATA FOR THE SECONDARY COLUMN (N=8)
Peak Number
(Figure 2.)
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
Analyte
1,2 dibromopropane (IS)
Formaldehyde
E-Acetaldehyde
Z-Acetaldehyde
E-Propanal
Z-Propanal
E-Butanal
Z-Butanal
E-Pentanal
Z-Pentanal
E-Hexanal
Z-Hexanal
Cyclohexanone
E-Heptanal
Z-Heptanal
E-Octanal
Z-Octanal
2,4,5, trifluoroacetophenone (S)
Benzaldehyde
E-Nonanal
Z-Nonanal
Decanal
Glyoxal
Methyl Glyoxal
Average
Retention
Time (min)
0.808
1.29
1.69
1.73
2.07
2.10
2.48
2.52
2.92
2.96
3.37
3.40
3.76
3.80
3.82
4.22
4.23
4.36
4.53
4.62
4.63
5.01
5.57
5.68
Standard
Deviation
1.48E-03
1.75E-03
1.87E-03
1.79E-03
1.79E-03
1.87E-03
1.76E-03
1.85E-03
1.56E-03
1.58E-03
1.59E-03
1.53E-03
1.43E-03
1.46E-03
1.48E-03
1.14E-03
7.25E-04
1.09E-03
1.29E-03
1.28E-03
1.31E-03
1.07E-03
9.91E-04
7.85E-04
Relative
Standard
Deviation
0.18%
0.14%
0.11%
0.10%
0.09%
0.09%
0.07%
0.07%
0.05%
0.05%
0.05%
0.05%
0.04%
0.04%
0.04%
0.03%
0.02%
0.03%
0.03%
0.03%
0.03%
0.02%
0.02%
0.01%
Secondary Column:
           AT-1701, 10 m x 0.10 mm i.d., 0.10 |j,m film thickness, injector temp. 200 °C, liner
           2 mm with a central 2 cm silanized glass wool plug, injection volume 1 (jL, split
           ratio 30:1, constant head pressure @ 32 psi, detector temp. 300 °C, detector make up
           flow 20 mL/minute. Temperature program: 70 °C initial, program at 27 °C/minute to
           220 °C, ballistic heating to 280 °C for burnout and hold at 280 °C for 0.4 minutes.
           Data collection via HP GC Chemstation at a rate of 50 Hz.

Carrier gas: Hydrogen (UHP)

Detector gas: 95:5 Argon Methane
                                     27

-------
 TABLE 3.  PRECISION AND ACCURACY IN REAGENT WATER (N=8)
                          5 |ug/L Fortification
Analyte
Formaldehyde
Acetaldehyde
Propanal
Butanal
Pentanal
Hexanal
Cyclohexanone
Heptanal
Octanal
Benzaldehyde
Nonanal
Decanal
Glyoxal
Methyl Glyoxal
Fortified
Concentration
(Hg/L)
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
Average
Concentration
(HR/L)
6.01
4.46
6.61
5.54
5.54
5.55
6.19
6.77
5.22
4.50
5.42
5.47
4.60
4.52
Unfortified
Sample
(HR/L)
0.303
ND
1.46
ND
ND
ND
ND
1.68
ND
ND
0.403
ND
ND
ND
Relative
Standard
Deviation
1.9%
2.5%
2.0%
2.4%
2.9%
4.0%
3.2%
5.0%
5.4%
4.1%
4.6%
5.1%
6.3%
5.7%
Average
Percent
RecoveryT
114%
89%
103%
111%
111%
111%
124%
102%
104%
90%
100%
109%
92%
90%
                          20 |ug/L Fortification
Analyte
Formaldehyde
Acetaldehyde
Propanal
Butanal
Pentanal
Hexanal
Cyclohexanone
Heptanal
Octanal
Benzaldehyde
Nonanal
Decanal
Glyoxal
Methyl Glyoxal
Fortified
Concentration
(Hg/L)
20.0
20.0
20.0
20.0
20.0
20.0
20.0
20.0
20.0
20.0
20.0
20.0
20.0
20.0
Average
Concentration
(HR/L)
20.8
20.7
20.6
20.0
20.2
20.3
20.7
20.8
20.2
19.6
20.3
20.3
19.5
19.3
Unfortified
Sample
(HR/L)
0.303
ND
1.46
ND
ND
ND
ND
1.68
ND
ND
0.403
ND
ND
ND
Relative
Standard
Deviation
2.3%
6.3%
2.4%
2.6%
3.0%
3.2%
2.0%
3.1%
2.2%
2.7%
2.6%
3.2%
4.1%
4.0%
Average
Percent
RecoveryT
102%
104%
96%
100%
101%
101%
104%
95%
101%
98%
100%
101%
98%
97%
t These recovery values were calculated using the equation in Section 9.7.2.
                                28

-------
TABLE 4. METHOD DETECTION LIMITS IN REAGENT WATER (n = 7)
Analyte
Formaldehyde
Acetaldehyde
Propanal
Butanal
Pentanal
Hexanal
Cyclohexanone
Heptanal
Octanal
Benzaldehyde
Nonanal
Decanal
Glyoxal
Methyl Glyoxal
Fortified
Concentration
(Hg/L)
1.0
1.0
1.0
1.0
1.0
1.0
1.0
1.0
1.0
1.0
1.0
1.0
1.0
1.0
Primary
Column
MDL (ng/L)
0.09
0.18
0.11
0.09
0.09
0.10
0.19
0.40
0.22
0.19
0.62
0.46
0.39
0.26
Secondary
Column
MDL (ng/L)
0.08
0.12
0.06
0.06
0.06
0.04
0.09
0.24
0.84
0.04
0.64
0.35
0.13
0.12
                         29

-------
TABLE 5. PRECISION AND ACCURACY IN CHLORINATED SURFACE
           WATER (N=7)

                            5 |ug/L Fortification
Analyte
Formaldehyde
Acetaldehyde*
Propanal
Butanal
Pentanal
Hexanal
Cyclohexanone
Heptanal
Octanal
Benzaldehyde
Nonanal
Decanal
Glyoxal
Methyl Glyoxal
Fortified
Concentration
(Hg/L)
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
Average
Concentration
(WS/L)
8.45
6.53
5.68
5.73
5.43
5.48
6.02
5.64
4.84
4.92
5.25
5.78
7.92
6.42
Unfortified
Sample
(WS/L)
3.40
1.76
0.620
0.390
ND
ND
0.650
0.840
ND
ND
0.250
ND
1.40
0.380
Relative
Standard
Deviation
3.3%
2.7%
2.2%
2.4%
2.6%
2.8%
4.2%
4.1%
6.4%
3.1%
8.5%
8.9%
9.2%
9.2%
Average
Percent
RecoveryT
101%
96%
101%
107%
109%
110%
107%
96%
97%
98%
100%
116%
130%
121%
                           20 |ug/L Fortification
Analyte
Formaldehyde
Acetaldehyde*
Propanal
Butanal
Pentanal
Hexanal
Cyclohexanone
Heptanal
Octanal
Benzaldehyde
Nonanal
Decanal
Glyoxal
Methyl Glyoxal
Fortified
Concentration
(Hg/L)
20.0
20.0
20.0
20.0
20.0
20.0
20.0
20.0
20.0
20.0
20.0
20.0
20.0
20.0
Average
Concentration
(WS/L)
22.6
20.1
20.4
20.1
20.5
20.7
20.5
19.1
18.8
20.4
19.9
20.8
25.9
23.0
Unfortified
Sample
(WS/L)
3.40
1.76
0.620
0.390
ND
ND
0.650
0.840
ND
ND
0.250
ND
1.40
0.380
Relative
Standard
Deviation
1.1%
1.8%
1.3%
1.8%
1.9%
2.2%
2.1%
4.1%
7.7%
2.2%
11.2%
10.5%
6.5%
10.3%
Average
Percent
RecoveryT
96%
91%
99%
99%
103%
103%
99%
91%
94%
102%
98%
104%
122%
113%
*Data for acetaldehyde were taken from the secondary column due to an interference with E-
 acetaldehyde.
t These recovery values were calculated using the equation in Section 9.7.2.
                                  30

-------
 TABLE 6 . PRECISION AND ACCURACY IN CHLORINATED GROUND
            WATER (N=7)

                            5 |ug/L Fortification
Analyte
Formaldehyde
Acetaldehyde*
Propanal
Butanal
Pentanal
Hexanal
Cyclohexanone
Heptanal
Octanal
Benzaldehyde
Nonanal
Decanal
Glyoxal
Methyl Glyoxal
Fortified
Concentration
(HR/L)
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
Mean
Concentration
(HR/L)
7.10
6.13
5.56
5.58
5.36
5.15
6.02
5.40
4.90
4.60
5.02
5.29
5.82
4.94
Unfortified
Sample
(HR/L)
2.21
1.13
0.657
0.437
ND
0.120
0.534
0.883
ND
ND
ND
ND
0.471
0.202
Relative
Standard
Deviation
1.3%
3.8%
3.1%
3.5%
3.5%
3.7%
4.7%
7.0%
7.9%
5.3%
8.7%
10.4%
10.9%
10.7%
Average
Percent
RecoveryT
97.8%
101%
98.0%
103%
107%
101%
110%
90.3%
98.0%
92.1%
100%
106%
107%
94.7%
                            20 |ug/L Fortification
Analyte
Formaldehyde
Acetaldehyde*
Propanal
Butanal
Pentanal
Hexanal
Cyclohexanone
Heptanal
Octanal
Benzaldehyde
Nonanal
Decanal
Glyoxal
Methyl Glyoxal
Fortified
Concentration
(Hg/L)
20.0
20.0
20.0
20.0
20.0
20.0
20.0
20.0
20.0
20.0
20.0
20.0
20.0
20.0
Mean
Concentration
(WS/L)
21.3
20.2
19.9
20.3
20.2
19.7
20.3
19.7
20.5
19.6
20.5
20.7
22.3
19.9
Unfortified
Sample
(WS/L)
2.21
1.13
0.657
0.437
ND
0.120
0.534
0.883
ND
ND
ND
ND
0.471
0.202
Relative
Standard
Deviation
2.8%
4.0%
3.3%
3.0%
3.7%
5.8%
4.8%
5.5%
6.2%
4.7%
6.6%
8.9%
10.3%
8.9%
Average
Percent
RecoveryT
95%
96%
96%
99%
101%
98%
99%
94%
102%
98%
103%
104%
109%
98%
* Data for acetaldehyde were taken from the secondary column due to an interference with E-
 acetaldehyde.
t These recovery values were calculated using the equation in Section 9.7.2.
                                  31

-------
TABLE 7. HOLDING TIME DATA FOR SAMPLES FROM AN
          UNTREATED SURFACE WATER SOURCE, FORTIFIED WITH
          METHOD ANALYTES AT 20 jug/L, WITH AND WITHOUT
          COPPER SULFATE BIOCIDE*

ANALYTE
Formaldehyde
Acetaldehyde
Propanal
Butanal
Pentanal
Hexanal
Cyclohexanone
Heptanal
Octanal
Benzaldehyde
Nonanal
Decanal
Glyoxal
Methyl glyoxal
% Recovery without Copper
Sulfate
DayO
104
96
94
92
87
83
94
83
82
94
72
50
103
108
Day 6
144
23
21
20
19
21
99
20
18
83
15
<10
98
68
Day 14
569
21
19
18
16
17
84
16
<10
74
<10
<10
37
<10
Day 21
619
24
22
21
21
22
78
17
11
79
<10
<10
<10
<10
% Recovery with Copper Sulfate
DayO
105
98
98
99
96
93
96
96
99
98
104
107
106
111
Day 6
105
99
98
98
98
97
101
94
96
100
98
101
108
105
Day 14
109
103
103
102
100
100
98
97
96
104
92
93
106
94
Day 21
106
98
96
91
94
92
94
91
93
92
84
82
90
73
* These data were collected and presented as Table 6 of Method 556, revision 1.0.
- All samples were stored headspace free at 4 °C.
- Values at all time points are the mean of  5 replicate analyses.  RSDs for replicate analyses of
samples containing copper sulfate were <10%. RSDs for unpreserved samples were higher due to
the degradation process.
                                  32

-------
TABLE 8. INITIAL DEMONSTRATION OF CAPABILITY
        REQUIREMENTS
Reference
Section
9.3
Section
9.2.2
Section
9.2.3
Section
9.2.4
Section
9.2.5
Requirement
Initial
Demonstration of
Low System
Background
Initial
Demonstration of
Precision (IDP)
Initial
Demonstration of
Accuracy
Method Detection
Limit (MDL)
Determination
Minimum
Reporting Levels
(MRLs)
Specification and Frequency
Analyze method blank and
determine that all target analytes
are below 1/2 the MRL
prior to performing IDC
Analyze 4-7 replicate LRBs
fortified at 20.0 g/L(ormid
cal.) on at least 2 different days
Calculate average recovery of
IDP replicates
a) select a fortifying level at 2 -
5 x the noise level
b) analyze 7 replicates in
reagent water taken thru all
steps
c) calculate MDL via equation -
do not subtract blank
d) replicate extractions and
analyses must be conducted
over at least 3 days
MRLs should be established for
all analytes during IDC, and be
updated as additional LRB data
is available.
Acceptance Criteria
The LRB concentration
must be < 1/2 of the
intended MRL
RSDmustbe < 20 %
Mean recovery ± 20% of
true value

Establish the MRL for
each analyte, as the LRB
concentration + 3 a or 3
times the mean LRB
concentration, whichever
is greater.
                         33

-------
TABLE 9. QUALITY CONTROL REQUIREMENTS (SUMMARY)
Reference
Section
10.2








Section
9.3



Section
10.3.1




Section
10.3.2




Section
8.4
Section
9.5


Section
9.6

Section
9.7


Requirement
Initial
Calibration








Laboratory
Reagent Blank
(LRB)


Continuing
Calibration
Check (CCC)
Option


Daily
Calibration
Option



Field Reagent
Blanks (FRB)
Internal
Standard (IS)


Surrogate
Standard
(SUR)
Laboratory
Fortified
Sample Matrix
(LFM)
Specification and Frequency
Use internal standard technique to
generate curve with five standards
that span the approximate range
of 5-40 ng/L. First or second
order curves must be forced
through zero. Either sum E/Z
isomer areas or average the
amount of each isomer.
Calculate E/Z ratios for analytes.
Run QCS.
Include LRB with each extraction
batch (up to 20 samples).
Analyze prior to analyzing
samples and determine to be free
of interferences.
Verify initial calibration by
running CCCs prior to analyzing
samples, after 10 samples, and
after the last sample.


Calibrate daily, but verify that
sensitivity and performance have
not changed significantly since
IDC.


1 per shipping batch

1,2-Dibromopropane is added to
all samples, blanks and standards


2', 4', 5' -Trifluoroacetophenone is
added samples, blanks and
standards
Fortify at least one sample per
analysis batch (20 samples or
less) at a concentration close to
that in the native sample.
Acceptance Criteria


QCS must agree within
70-130%.

Lowest concentration
should be near MRL.



All analytes < 1/2 MRL



Recovery for mid-level
CCC must be 70- 130%
of the true value,
recovery for low level
must be 50-150% of the
true value.
Peak areas for IS, SUR,
and method analytes for
mid-level CAL std must
be +/- 50% of the peak
areas obtained for that
CAL std during IDC.
All analytes < 1/2 MRL

IS area counts must be 70
- 130% of the average
initial calibration area
counts
Surrogate recovery must
be 70- 130% of the true
value.
Recoveries not within 70-
130% may indicate
matrix effect

                          34

-------
Section
  9.8
Field
Duplicates
Extract and analyze at least one
duplicate sample with every
analysis batch (20 samples or
less)	
Suggested RPD < 30 %
Section
  9.9
Quality
Control
Sample (QCS)
Analyze a QCS at least quarterly
from an external/second source.
QCS must agree within
70-130%.
Section
 9.10,
Section
 10.2.5
  and
Section
  12.4
E/Z Isomer
Ratio
Agreement
Calculate the E/Z isomer ratio for
target analytes and compare to
E/Z ratio in initial calibration
E/Z ratio in standards,
blanks, and samples must
be within ±50% of E/Z
ratio in initial calibration.

Do not report value if one
isomer is missing.	
Section
 8.3.1
Sample
Holding Time
Properly preserved samples may
be stored in the dark at 4 °C for 7
days.
Do not report data for
samples that have
exceeded their holding
time, or that have not
been properly preserved
or stored.
Section
 8.3.2
Extract
Holding Time
Extracts may be stored in the dark
at 4 °C for 14 days.
Do not report data for
extracts that have
exceeded their holding
time.
                                     35

-------
Hz
2000
1750
1500

1250
1000
750 -
500
250
0
*CN *





T


— — '



IjuJ








L — 1
FIGURE 1.
> Primary Column Chromatogram (DB-5) *«



CO



I

u




•>
t*
CO



OO


^~







1
u





c


L
D
0
^^ T
T


LA 	 *_J
p
T
CN


LAJ
0
~ I
i
^


I
0
^
sf


VJl











CS.CO



o>
00
CD
i

L^

LJ


U
<
<


I — J
^ s


1 	 1




s






0.5         1           1.5          2



  * Peaks have been attenuated
            2.5
3.5
4.5
                                                                                mm
36

-------
                                           FIGURE 2.
Hz
3000
2500
2000 n
1500
1000
500
0
*


T


^ 	 ^
S€
M
*

f



T

vj
;condary Column Chromatogram (AT- 1701)
*

IS

IL
5
1-
CD

••
CO



<
I
J>
0
T— T1
T1
[ 	 |
oo
CM
IL,
^-A^J \__/LJ




in
CD
U
O
T
Wl
0
L^_
T
C
<


M
3 j
i
C
T
C
C


M
M
A.

5
o"
M


I
0
N




0.5         1                 2
       * Peaks have been attenuated
mm
                               37

-------
THIS PAGE LEFT INTENTIONALLY BLANK
 38

-------