EPA Document #: EPA/600/R-05/053
METHOD 535.    MEASUREMENT OF CHLOROACETANILIDE AND OTHER
                ACETAMIDE HERBICIDE DEGRADATES IN DRINKING
                WATER BY SOLID PHASE EXTRACTION AND LIQUID
                CHROMATOGRAPHY/TANDEM MASS SPECTROMETRY
                (LC/MS/MS)
                               Version 1.1
                               April 2005
                             J.A. Shoemaker
                              M.V. Bassett
             NATIONAL EXPOSURE RESEARCH LABORATORY
                OFFICE OF RESEARCH AND DEVELOPMENT
              U. S. ENVIRONMENTAL PROTECTION AGENCY
                        CINCINNATI, OHIO 45268


                                 535-1

-------
                                  METHOD 535
1.
MEASUREMENT OF CHLOROACETANILIDE AND OTHER ACETAMIDE
  HERBICIDE DEGRADATES IN DRINKING WATER BY SOLID PHASE
   EXTRACTION AND LIQUID CHROMATOGRAPHY/TANDEM MASS
                     SPECTROMETRY (LC/MS/MS)

 SCOPE AND APPLICATION
      1.1   This is a liquid chromatographic (LC)/tandem mass spectrometric (MS/MS)
            method applicable to the determination of the ethanesulfonic acid (ESA) and
            oxanilic acid (OA) degradates of the chloroacetanilide and other acetamide
            herbicides in finished drinking waters.  The single laboratory Lowest
            Concentration Minimum Reporting Level (LCMRL) has also been determined in
            reagent water.1 The following compounds can be determined using this method:
Analyte
Acetochlor ESA
Acetochlor OA
Alachlor ESA
Alachlor OA
Dimethenamid ESA
Dimethenamid OA
Flufenacet ESA
Flufenacet OA
Metolachlor ESA
Metolachlor OA
Propachlor ESA
Propachlor OA
Chemical Abstracts Service (CAS)
Registry Number
187022-11-3
184992-44-4
142363-53-9
171262-17-2
205939-58-8
—
—
—
171118-09-5
152019-73-3
—
—
      1.2    The Minimum Reporting Level (MRL) is the lowest analyte concentration that
             meets Data Quality Objectives (DQOs) that are developed based on the intended
                                      535-2

-------
             use of this method.  The lowest concentration MRL (LCMRL) is a single
             laboratory determination of the lowest true concentration for which a future
             recovery is expected, with 99% confidence, to be between 50 and 150% recovery.
             The procedure used to determine the LCMRL is described elsewhere.1  Single
             laboratory LCMRLs for analytes in this method range from 0.016-0.11 ug/L for
             the ion trap instrument and 0.22 - 0.55 ug/L for the triple quadrupole instrument,
             and are listed in Table 5.

       1.3    Laboratories using this method are not required to determine the LCMRL for this
             method, but must demonstrate that their laboratory MRL for this method meets
             the requirements described in Section 9.2.4.

       1.4    Detection limit (DL) is defined as the statistically calculated minimum
             concentration that can be measured with 99% confidence that the reported value is
             greater than zero.2 The DL is compound dependent and is dependent on
             extraction efficiency, sample matrix, fortification concentration, and instrument
             performance. Determining the DL for analytes in this method is optional
             (Sect. 9.2.5).  DLs for target analytes range from 0.0060 - 0.043 ug/L for the ion
             trap instrument and 0.057 -0.11 ug/L for the triple quadrupole instrument, and are
             listed in Table 5.

       1.5    This method is intended for use by or under the supervision of analysts  skilled in
             solid phase extractions, LC/MS/MS analysis and the interpretation of LC/MS/MS
             chromatograms and mass spectra.

2.     SUMMARY OF METHOD

       2.1    Analytes and surrogate are extracted and concentrated by passing a 250-mL water
             sample through a solid phase extraction (SPE) cartridge containing 0.5 g of
             nonporous  graphitized carbon.  The compounds are eluted from the solid phase
             with a small quantity of methanol containing 10 mM ammonium acetate.  The
             methanol extract is then concentrated to dryness and reconstituted with  1 mL of
             water containing 5 mM ammonium acetate. The analytes are chromatographically
             separated by injecting an aqueous aliquot (100 uL) into an LC system equipped
             with a reversed phase (Clg) column.  After elution from the column, the analytes
             are detected by liquid chromatography/tandem mass spectrometry (LC/MS/MS).
             The concentration of each identified  component is measured by an internal
             standard procedure, i.e. relating the product ion response of that compound to the
             product ion response of the compound that is used as an internal standard. A
             surrogate analyte, whose concentration is known in every sample, is measured
             with the same internal standard calibration procedure.
                                         535-3

-------
3.      DEFINITIONS

       3.1    ANALYSIS BATCH - A set of samples that is analyzed on the same instrument
             during a 24-hour period that begins and ends with the analysis of the appropriate
             Continuing Calibration Check (CCC) standards.  Additional CCCs may be
             required depending on the length of the analysis batch and/or the number of field
             samples.

       3.2    CALIBRATION STANDARD (CAL) - A solution prepared from the primary
             dilution standard solution and/or stock standard solution, internal standard(s), and
             the surrogate(s).  The CAL solutions are used to calibrate the instrument response
             with respect to analyte concentration.

       3.3    COLLISIONALLY ACTIVATED DISSOCIATION (CAD) -  The process of
             converting the precursor ion's translational energy into internal energy by
             collisions with neutral gas molecules to bring about dissociation into product ions.

       3.4    CONTINUING CALIBRATION CHECK (CCC) - A  calibration standard
             containing the target analytes, internal standard(s) and surrogate(s), which is
             analyzed periodically to verify the accuracy of the existing calibration for those
             analytes.

       3.5    DETECTION LIMIT (DL) - The minimum concentration of an analyte that can
             be identified, measured and reported with 99% confidence that the  analyte
             concentration is greater than zero. This is a statistical  determination of precision
             (Sect. 9.2.5), and accurate quantitation is not expected at this level.2

       3.6    EXTRACTION BATCH - A set of up to 20 field samples (not including QC
             samples) extracted together by the same person(s) during a work day using the
             same lot of solid phase extraction devices, solvents, surrogate, internal standard
             and fortifying solutions. Required QC samples include Laboratory Reagent
             Blank, Laboratory Fortified Blank, Laboratory Fortified Sample Matrix,  and either
             a Field Duplicate or Laboratory Fortified Sample Matrix Duplicate.

       3.7    FIELD DUPLICATES (FD1 and FD2) - Two separate samples collected at the
             same time and place under identical circumstances, and treated exactly the same
             throughout field and laboratory procedures.  Analyses  of FD1 and FD2 give a
             measure of the precision associated with sample collection, preservation, and
             storage, as well as laboratory procedures.

       3.8    INTERNAL STANDARD (IS) - A pure analyte(s) added to a  sample, extract, or
             standard solution in known amount(s) and used to measure the relative responses
                                         535-4

-------
       of other target analytes and surrogates that are components of the same solution.
       The internal standard must be an analyte that is not a sample component.

3.9    LABORATORY FORTIFIED BLANK (LFB) - An aliquot of reagent water or
       other blank matrix to which known quantities of the target analytes and all the
       preservation compounds are added in the laboratory.  The LFB is analyzed exactly
       like a sample, and its purpose is to determine whether the methodology is in
       control, and whether the laboratory is capable of making accurate and precise
       measurements.

3.10   LABORATORY FORTIFIED SAMPLE MATRIX (LFSM) - An aliquot of a
       preserved field sample to which known quantities of the target analytes are added
       in the laboratory. The LFSM is processed and analyzed exactly like a sample, and
       its purpose is to determine whether the sample matrix contributes bias to the
       analytical results. The background concentrations of the analytes in the sample
       matrix must be determined in a separate sample extraction and the measured
       values in the LFSM corrected for background concentrations.

3.11   LABORATORY FORTIFIED SAMPLE MATRIX DUPLICATE (LFSMD) - A
       second aliquot of the field sample used to prepare the LFSM. The LFSMD is
       fortified, extracted, and analyzed identically to the LFSM. The LFSMD is used
       instead of the Field Duplicate to assess method precision when the occurrence of
       target analytes are low.

3.12   LABORATORY REAGENT BLANK (LRB) - An aliquot of reagent water or
       other blank matrix that is treated exactly as a sample including exposure to all
       glassware, equipment, solvents and reagents, sample preservatives, internal
       standard, and surrogates that are used in the analysis batch.  The LRB is used to
       determine if target analytes or other interferences are present in the laboratory
       environment, the reagents, or the apparatus.

3.13   LOWEST CONCENTRATION MINIMUM REPORTING LEVEL (LCMRL) -
       The single laboratory LCMRL is the lowest true concentration for which a future
       recovery is expected, with 99% confidence, to be between 50 and 150% recovery.1

3.14   MATERIAL SAFETY DATA SHEET (MSDS) - Written information provided
       by vendors concerning a chemical's toxicity, health hazards, physical properties,
       fire, and reactivity data including storage, spill, and handling precautions.

3.15   MINIMUM REPORTING LEVEL (MRL) - The minimum concentration that can
       be reported as a quantitated value for a target analyte in a sample following
       analysis. This defined concentration can be no lower than the concentration of the
       lowest calibration standard for that analyte and can only be used if acceptable
                                 535-5

-------
             quality control criteria for this standard are met.  A procedure for verifying a
             laboratory's MRL is provided in Section 9.2.4.

       3.16   PRECURSOR ION - For the purpose of this method, the precursor ion is the
             deprotonated molecule ([M-H]") of the target analyte. In MS/MS, the precursor
             ion is mass selected and fragmented by collisionally activated dissociation to
             produce distinctive product ions of smaller m/z.

       3.17   PRIMARY DILUTION STANDARD (PDS) SOLUTION - A solution containing
             the analytes prepared in the laboratory from stock standard solutions and diluted
             as needed to prepare calibration solutions and other needed analyte solutions.

       3.18   PRODUCT ION - For the purpose of this method, a product ion is one of the
             fragment ions produced in MS/MS by collisionally activated dissociation of the
             precursor ion.

       3.19   QUALITY CONTROL SAMPLE (QCS) - A solution of target analytes of known
             concentrations that is obtained from a source external to the laboratory and
             different from the source of calibration standards. It is used to check laboratory
             performance with externally prepared test materials.

       3.20   STOCK STANDARD SOLUTION (SSS) - A concentrated solution containing
             one or more target analytes prepared in the laboratory using assayed reference
             materials or purchased from a reputable commercial source.

       3.21   SURROGATE ANALYTE (SUR) - A pure analyte, which chemically resembles
             target analytes and is extremely unlikely to be found in any sample.  This analyte
             is added to a sample aliquot in known amount(s) before processing and is
             measured with the same procedures used to measure other sample components.
             The purpose of the SUR is to monitor method performance with each sample.

4.      INTERFERENCES

       4.1    Interfering contamination may occur when a sample containing low
             concentrations of compounds is analyzed immediately after a sample containing
             relatively high concentrations  of compounds.  After analysis of a sample
             containing high concentrations of compounds, a laboratory reagent blank should
             be analyzed to ensure that accurate values are obtained for the next sample.  In the
             case of automated analysis, the analyst may not be aware of high concentration
             samples until after an entire batch is analyzed. In this situation, the analyst should
             carefully review data from samples analyzed immediately after high concentration
             samples,  and reanalyze them if necessary.
                                         535-6

-------
       4.2    Matrix interferences may be caused by contaminants that are co-extracted from
             the sample. The extent of matrix interferences will vary considerably from source
             to source, depending upon the nature of the water.  Humic and/or fulvic material
             is co-extracted by this method and high levels can cause enhancement and/or
             suppression in the electrospray ionization source or low recoveries on the carbon
             SPE. For this method, total organic  carbon (TOC) is a good indicator of humic
             content of the sample. See Section 13.4 for more details.

       4.3    The carbon solid phase cartridges did not produce any interferences during
             method development. However, brands and lots of carbon cartridges should be
             tested to ensure that contamination does not preclude analyte identification and
             quantitation. The analysis of laboratory reagent blanks can provide important
             information regarding the presence or absence of such interferences.

       4.4    Method interferences may be caused by contaminants in solvents, reagents
             (including reagent water), sample bottles and caps, and other sample processing
             hardware that lead to discrete artifacts and/or elevated baselines in the
             chromatograms. All items such as these must be routinely demonstrated to be free
             from interferences (less than 1/3 the MRL for each target analyte), under the
             conditions of the analysis, by analyzing laboratory reagent blanks as described in
             Section 9.3.  Subtracting blank values from sample results is not permitted.

5.     SAFETY

       5.1    The toxicity or carcinogenicity of each reagent used in this method has not been
             precisely defined; each chemical compound should be treated as a potential health
             hazard, and exposure to these chemicals should be minimized. The laboratory is
             responsible for maintaining a current awareness file of OSHA regulations
             regarding the safe handling of the chemicals specified in this method. A reference
             file of MSDSs should also be made available to all personnel involved in the
             chemical analysis.

       5.2    Pure standard materials and stock standard solutions of these compounds should
             be handled with suitable protection to skin, eyes, etc.3"5

6.     EQUIPMENT AND SUPPLIES (All specifications are suggested. Brand names and/or
       catalog numbers are included for illustration only, and do not imply endorsement of the
       product.)

       6.1    GLASSWARE - All glassware must be meticulously cleaned. Wash glassware
             with detergent and tap water, rinse with tap water, followed by reagent water.  A
             final rinse with solvents may be needed.  In place of a solvent rinse, non-
                                         535-7

-------
      volumetric glassware can be heated in a muffle furnace at 400 °C for 2 hours.
      Volumetric glassware should not be heated above 120 °C.

6.2   SAMPLE CONTAINERS - Amber glass bottles (250 mL or larger) fitted with
      PTFE (polytetrafluoroethylene) lined screw caps.

6.3   VIALS - Amber glass autosampler vials with PTFE faced septa for storing
      standards and extracts.

6.4   VOLUMETRIC FLASKS - Class A, various sizes used for preparation of
      standards.

6.5   GRADUATED CYLINDERS - Various sizes.

6.6   MICRO SYRINGES - Various sizes.

6.7   BALANCE - Analytical, capable of accurately weighing to 0.0001 g.

6.8   DISPOSABLE PIPETTES - 1 mL, used to transfer sample extracts to
      autosampler vials.

6.9   CONICAL COLLECTION TUBES -15 mL, or other glassware suitable for
      collection of the eluent from the solid phase cartridge after extraction.

6.10  SOLID PHASE EXTRACTION (SPE) APPARATUS USING CARTRIDGES

      6.10.1  SPE CARTRIDGES - 0.5 g, 6 mL SPE cartridges containing a nonporous
             graphitized carbon sorbent phase (Supelclean ENVI-Carb SPE cartridges,
             Supelco Catalog No. 57904 or equivalent).

      6.10.2  VACUUM EXTRACTION MANIFOLD - A manual vacuum manifold
             (Supelco Cat. No. 57030 and 57275 or equivalent) for cartridge
             extractions, or an automatic/robotic sample preparation (Zymark Autotrace
             SPE Workstation or equivalent) system designed for use with solid phase
             extraction cartridges, may be used if all quality control requirements
             discussed in Section 9 are met.  Extraction and/or elution steps may not be
             changed or omitted to accommodate the use of an automated system.

      6.10.3  SAMPLE DELIVERY SYSTEM - Use of a transfer tube system (Supelco
             "Visiprep"; cat. #: 57275 or equivalent), which transfers the sample
             directly from the sample container to the  SPE cartridge, is recommended.
             Sample reservoirs, which attach to the cartridge, may be used, although
             they hold only 75 mL at one time.
                                 535-8

-------
       6.10.4 LABORATORY OR ASPIRATOR VACUUM SYSTEM - Sufficient
             capacity to maintain a vacuum of approximately 25 cm (10 in) of mercury.

6.11    EXTRACT CONCENTRATION SYSTEM - Extracts are concentrated by
       blowdown with nitrogen using a water bath set at 60-65 °C (Meyer N-EVAP,
       model 111, Organomation Associates, Inc.  or equivalent). Other types of
       evaporation/concentration equipment may be used as long as all QC requirements
       in Section 9 are met.

6.12    LIQUID CHROMATOGRAPHY (LC)/TANDEM MASS
       SPECTROMETER(MS/MS)/DATA SYSTEM

       6.12.1 LC SYSTEM - Capable of reproducibly injecting up to 100-uL aliquots,
             and performing binary linear gradients at a constant flow rate near the flow
             rate used for development of this method, e.g., 0.4 mL/min. The use of a
             column heater is required. During method development, the column was
             heated to 65 °C to achieve adequate separation of alachlor ESA and
             acetochlor ESA, although other temperatures may be used as long as the
             resolution requirement in Section 9.10 is met. The LC must be capable of
             pumping the water/methanol mobile phase without the use of a vacuum
             degasser. Vacuum degassers will volatilize the ammonium acetate mobile
             phase causing the analyte peaks to shift to earlier retention times over the
             course of the analysis batch. Sparging the mobile phase prior to use for a
             short time period (approximately 5 minutes) is acceptable.

       6.12.2 LC/Tandem Mass Spectrometer - The LC/MS/MS must be capable of
             negative ion electrospray ionization (ESI) near the suggested LC flow rate
             of 0.4 mL/min. The system must be capable of performing MS/MS to
             produce unique product ions (Sect.  3.18) for the target analytes within
             specified retention time segments.  A minimum of 20 scans across the
             chromatographic peak is recommended to ensure adequate precision. Data
             is demonstrated in Tables 5-9 using an ion trap mass spectrometer
             (Finnigan LCQ Deca) and a triple quadrupole mass spectrometer
             (Micromass Quattro Micro).

       6.12.3 An interfaced data system is required to acquire, store, reduce, and output
             mass spectral data.  The computer software should have the capability of
             processing stored LC/MS/MS data by recognizing an LC peak within any
             given retention time window.  The software must allow integration of the
             ion abundance  of any specific ion within specified time or scan number
             limits. The software must be able to calculate relative response factors,
             construct linear regressions or quadratic calibration curves, and calculate
             analyte concentrations.
                                 535-9

-------
             6.12.4 ANALYTICAL COLUMN- An LC Clg column (2.1 x 100 mm) packed
                   with 5 um dp Clg solid phase particles (Agilent # : 79916OD-552 or
                   equivalent) was used on the triple quadrupole instrument. For the ion trap
                   instrument, an LC Clg column (2.1 x 200 mm) packed with 5 um dp Clg
                   solid phase particles (Agilent # : 79916OD-572 or equivalent) was used to
                   achieve more separation between target analytes. Any column that
                   provides adequate resolution, peak shape, capacity, accuracy, and
                   precision (Sect. 9) maybe used. The Clg columns employed in the method
                   development were used for over 10 months with no measurable
                   degradation due to the elevated column temperatures (60 to 70 °C).

7.     REAGENTS AND STANDARDS

      7.1    GASES, REAGENTS, AND SOLVENTS - Reagent grade or better chemicals
             should be used. Unless otherwise indicated, it is intended that all reagents shall
             conform to the specifications of the Committee on Analytical Reagents of the
             American Chemical Society, where such specifications are available.  Other
             grades may be used, provided it is first determined that the reagent is of
             sufficiently high purity to permit its use without lessening the quality of the
             determination.

             7.1.1  REAGENT WATER - Purified water which does not contain any
                   measurable quantities of any target analytes or interfering compounds
                   greater than 1/3 the MRL for each compound of interest.

             7.1.2  METHANOL (CH3OH, CAS#: 67-56-1) - High purity, demonstrated to be
                   free of analytes and interferences (Fisher Optima grade or equivalent).

             7.1.3  AMMONIUM ACETATE (NH4C2H3O2, CAS#: 631-61-8) -Highpurity,
                   demonstrated to be free of analytes and interferences (Sigma-Aldrich ACS
                   grade or equivalent).

             7.1.4  10 mM AMMONIUM ACETATE/METHANOL - To prepare 1 L, add
                   0.7708 g ammonium acetate to 1L of methanol.  During method
                   development, this solution was shown to be stable for at least 18 days.

             7.1.5  5 mM AMMONIUM ACETATE/REAGENT WATER - To prepare 1 L,
                   add 0.3854 g ammonium acetate to  1 L of reagent water (pH~6.5).  This
                   solution is prone to volatility losses and should be replaced at least every
                   48 hours.

             7.1.6   AMMONIUM CHLORIDE (NH4C1, CAS#: 12125-02-9) - Acts as a
                   sample dechlorinating agent (Sect. 8.1.1).
                                       535-10

-------
      7.1.7  NITROGEN - Used for the following purposes.

             7.1.7.1 Nitrogen aids in aerosol generation of the ESI liquid spray and
                    used as collision gas in some MS/MS instruments.  The nitrogen
                    used should meet or exceed instrument manufacturer's
                    specifications.

             7.1.7.2 Nitrogen is used to concentrate sample extracts, Ultra High Purity
                    or equivalent.

      7.1.8  ARGON (OPTIONAL) - Can be used as collision gas in some MS/MS
             instruments. Argon should meet or exceed instrument manufacturer's
             specifications.

7.2   STANDARD SOLUTIONS - Solution concentrations listed in this Section were
      used to develop this method and are included as an example. If compounds used
      to prepare solutions are 96% pure or greater, the weight may be used without
      correction for purity to calculate the concentration of the stock standard.
      Standards for sample fortification generally should be prepared in the smallest
      volume that can be accurately measured to minimize the addition of organic
      solvent to aqueous samples.  Even though stability times for standard
      solutions are suggested in the following sections, laboratories should use
      standard QC practices to determine when Standard Solutions described in
      this section need to be replaced.

      7.2.1  INTERNAL STANDARD ANALYTE (IS) STANDARD SOLUTION -
             BUTACHLOR ESA - This IS was not commercially available when the
             method was developed. Thus, the IS was synthesized by the procedure
             listed in Appendix A of this method. Commercially available standards
             may be used, if available.  Alternate internal standards may be used
             provided they meet the definition of an IS (Sect. 3.8) and they meet the QC
             requirements in Section 9.6.

             7.2.1.1 IS STOCK SOLUTION - If preparing from neat material,
                    accurately weigh approximately 5 mg of the neat IS to the nearest
                    0.1 mg into a tared, 1-mL volumetric flask. Dilute to the mark
                    with methanol. Stock solutions have been shown to be stable for
                    6 months when stored at 4 °C or less.

             7.2.1.2 IS PRIMARY DILUTION STANDARD (IS PDS) - Prepare  the
                    IS PDS by adding enough of the IS stock standard to a volumetric
                    flask partially  filled with methanol to make a final concentration
                    near 10 ug/mL when filled to the mark with methanol.  The IS
                                 535-11

-------
             PDS has been shown to be stable for 6 months when stored at 4 °C
             or less.

7.2.2  SURROGATE ANALYTE (SUR) STANDARD SOLUTION,
      DIMETHACHLOR ESA - This SUR was not commercially available
      when the method was developed. Thus, the SUR was synthesized by the
      procedure listed in Appendix B of this method.  Commercially available
      standards may be used, if available. Alternate surrogates may be used
      provided they meet the definition of a surrogate (Sect. 3.21) and they meet
      the QC requirements in Section 9.7.

      7.2.2.1 SUR STOCK SOLUTION - If preparing from neat material,
             accurately weigh approximately 5 mg of the neat SUR to the
             nearest 0.1 mg into a tared, 1-mL volumetric flask.  Dilute to the
             mark with methanol. Stock solutions have been shown to be stable
             for 6 months when stored at 4 °C or less.

      7.2.2.2 SUR PRIMARY DILUTION STANDARD (SUR PDS) - Prepare
             the SUR PDS by adding enough of the SUR stock standard to a
             volumetric flask partially filled with methanol to make a final
             concentration near 24 ug/mL when filled to the mark with
             methanol.  The SUR PDS has been shown to be stable for
             12 months when stored at 4 °C or less.

7.2.3  ANALYTE STANDARD SOLUTIONS - Obtain the analytes listed in the
      table in Section 1.1 as neat  standards. Prepare the Analyte  Stock and
      Primary Dilutions Standards as described below.

      7.2.3.1 ANALYTE STOCK STANDARD SOLUTION - If preparing
             from neat material, accurately weigh approximately 5 mg of pure
             material to the nearest 0.1 mg into a tared, 5-mL volumetric flask.
             Dilute to the mark with methanol. Repeat for each  target analyte.

      7.2.3.2 ANALYTE PRIMARY DILUTION STANDARD (PDS)
             SOLUTION (10-40 ug/mL) - The analyte PDS contains all the
             target analytes of interest in methanol at various concentrations.
             The ESI and MS/MS response varies by compound; therefore, a
             mix of concentrations may be needed in the analyte PDS. See
             Tables in Section 17 for suggested concentrations for each analyte.
             During method development, the analyte PDS was prepared such
             that approximately the same instrument response was obtained for
             all the analytes. The analyte PDS is prepared by dilution of the
             Analyte Stock Standard solution and is used to prepare the
                          535-12

-------
                          calibration standards, and fortify the LFBs, the LFSMs, the
                          LFSMDs and FDs with the target analytes. The analyte PDS has
                          been shown to be stable for 6 months when stored at 4 °C or less.

             7.2.4   CALIBRATION STANDARDS (CAL) - At least five calibration
                    concentrations are required to prepare the initial calibration curve spanning
                    a 20-fold concentration range (Sect. 10.2). Larger concentration ranges
                    will require more calibration points.  Prepare the calibration standards over
                    the concentration range of interest  from dilutions  of the analyte PDS in
                    5mM ammonium acetate/reagent water (Sect. 7.1.5). The target analyte
                    concentrations found in Tables 5-9 can be used as a starting point for
                    determining the calibration range.  The IS and SUR are added to the
                    calibration standards at a constant  concentration.  During method
                    development, the concentration of the SUR was 120 ug/L in the standard
                    (0.48 ug/L in the sample) and the IS was 80 ug/L. The lowest
                    concentration calibration standard  must be at or below the MRL, which
                    may depend on system sensitivity.  The CAL standards may be also be
                    used as CCCs. If stored, the aqueous standards must be stored refrigerated
                    in the same manner as the samples. During method development, the
                    calibration standards were shown to be stable for  at least two months.
                    Longer storage times are acceptable provided appropriate QC measures are
                    documented demonstrating the calibration standard stability.

8.      SAMPLE COLLECTION, PRESERVATION, AND STORAGE

       8.1    SAMPLE BOTTLE PREPARATION

             8.1.1   Prior to shipment to the field, ammonium chloride (Sect. 7.1.6) must be
                    added as a dry solid to each amber bottle fitted with a PTFE lined screw
                    cap. A 250-mL sample bottle or larger is recommended. Add ammonium
                    chloride to yield a sample concentration of 100 mg/L.  During method
                    development, 25 to 30 mg of ammonium chloride was added to each
                    250-mL sample bottle.

             8.1.2   Sample bottles must not be prerinsed with sample before collection.
                    Doing so will wash out the preservative added to the bottles prior to
                    shipment.
                                        535-13

-------
       8.2    SAMPLE COLLECTION

             8.2.1   Open the tap, and allow the system to flush until the water temperature has
                    stabilized (usually about 2 min).  Collect samples from the flowing
                    system.

             8.2.2   When sampling from an open body of water, fill a wide-mouth bottle or
                    beaker with sample from a representative area, and carefully fill sample
                    bottles from the container. Sampling equipment, including automatic
                    samplers, must be free of plastic tubing, gaskets, and other parts that may
                    leach interfering analytes into the water sample.

             8.2.3   Fill sample bottles, taking care not to flush out the sample preservation
                    reagents. Samples do not need to be collected headspace free.

             8.2.4   After collecting the sample, cap carefully to avoid spillage, and agitate by
                    hand for 1 minute.  Keep samples sealed from collection time until
                    analysis.

    8.3      SAMPLE SHIPMENT AND STORAGE - All samples should be cooled (ice or
             chemical freeze packs) during shipment and must not exceed 10 °C during the first
             48 hours after collection.  Samples must be confirmed to be at or  below 10 °C
             when they are received at the laboratory.  Samples stored in the lab must be held
             at or below 6 °C and protected from light until analysis.  Samples should not be
             frozen.

             NOTE: Samples that are significantly above 10 °C at the time of collection may
             need to be iced  or refrigerated for a period of time, in order to chill them prior to
             shipping.  This will allow them to be shipped with sufficient ice (or chemical
             freeze packs) to meet the above requirements.

    8.4      SAMPLE AND EXTRACT HOLDING TIMES  - Results of the sample storage
             stability study of all target analytes indicated that all compounds are stable for
             14 days in water samples that are collected, dechlorinated, shipped and stored as
             described in Sections 8.1-8.3.  Therefore, aqueous samples must be extracted
             within 14 days.   Sample extracts maybe stored at 4 °C or less for  up to 28 days
             after sample extraction. Data from holding time studies are shown in Tables 10
             and 11.

9.   QUALITY CONTROL

    9.1      QC requirements include the Initial Demonstration of Capability and ongoing QC
             requirements that must be met when preparing and analyzing field samples. This
                                        535-14

-------
         section describes each QC parameter, their required frequency, and the
         performance criteria that must be met in order to meet EPA quality objectives.
         The QC criteria discussed in the following sections are summarized in Section 17,
         Tables 12 & 13.  These QC requirements are considered the minimum acceptable
         QC criteria. Laboratories are encouraged to institute additional QC practices to
         meet their specific needs.

         9.1.1  METHOD MODIFICATIONS - The analyst is permitted to modify LC
               columns  and conditions provided the QC in Section 9 is still met.  The
               analyst is also allowed to modify the internal and surrogate standards.
               Each time such method modifications are made, the analyst must repeat
               the procedures of the IDC (Sect.  9.2). The analyst may not modify the LC
               mobile phase components or LC modifiers  (Sect. 13.4.1)

               Caution: Matrix enhancement is a significant  problem when
               analyzing SPE extracts by LC/MS/MS. Thus, if the analyst chooses to
               modify the LC conditions (e.g., flow rate, temperature, gradient), care
               must be  taken ensure that the target analytes do not elute too closely
               to the major portion of the TOC elution. An LC/MS analysis of a
               spiked matrix extract will be necessary to evaluate this.

9.2      INITIAL DEMONSTRATION OF CAPABILITY  (IDC) - The IDC must be
         successfully performed prior to analyzing any field samples. Prior to conducting
         the IDC, the analyst must first generate an acceptable Initial Calibration following
         the procedure outlined in Section 10.2.

         9.2.1  INITIAL DEMONSTRATION OF LOW SYSTEM BACKGROUND -
               Before any field samples are analyzed, and any time a new set of reagents
               or SPE cartridges is used, it must be demonstrated that a laboratory reagent
               blank is reasonably free of  contamination and that the criteria in
               Section 9.3 are met.

         9.2.2   INITIAL DEMONSTRATION OF PRECISION - Prepare and analyze
               4-7 replicate LFBs fortified at or near the mid-range of the initial
               calibration curve, according to the procedure described in Section 11.
               Ammonium chloride, as described in Section 8.1.1, must also be added to
               these samples. The relative standard deviation (RSD)  of the results of the
               replicate analyses must be <20% for all target analytes and the surrogate.

         9.2.3  INITIAL DEMONSTRATION OF  ACCURACY - Using the same set of
               replicate  LFB data generated for Section 9.2.2, calculate the average
               recovery. The average recovery of the replicate values must be within
               ±30% of the true value.
                                   535-15

-------
9.2.4   MINIMUM REPORTING LEVEL (MRL) CONFIRMATION - Establish
       a target concentration for the MRL based on the intended use of the
       method. Establish an Initial Calibration following the procedure outlined
       in Section 10.2.  The lowest calibration standard used to establish the
       Initial Calibration (as well as the low-level Continuing Calibration Check
       standard) must be at or below the concentration of the MRL.  Establishing
       the MRL concentration too low may cause repeated failure of ongoing QC
       requirements. Confirm or validate the MRL following the procedure
       outlined below.1

       9.2A.I  Fortify, extract, and analyze seven replicate Laboratory Fortified
               Blanks at the proposed MRL concentration.  These LFBs must
               contain ammonium chloride as described in Section 8.1.1.
               Calculate the mean (Mean) and standard deviation for these
               replicates.  Determine the Half Range for the prediction interval
               of results (HRPIR) using the equation below

                                HRPSL = 3.963S

               where S is the standard deviation, and 3.963 is a constant value
               for seven replicates.

       9.2.4.2  Confirm that the upper and lower limits for the Prediction Interval
               of Result (PIR = Mean ± HRPIR) meet the upper and lower
               recovery limits as shown below.

               The Upper PIR Limit must be < 150% recovery.
                               Mean + HRpm
                                                 *100 < 150%
                          For&jied Concentration


              The Lower PIR Limit must be > 50% recovery.


                                Mean - HRprs
                           Fortified Concentration
                                                  xl 00 > 50%
       9.2.4.3 The MRL is validated if both the Upper and Lower PIR Limits
              meet the criteria described above (Sect. 9.2.4.2). If these criteria
              are not met, the MRL has been set too low and must be
              determined again at a higher concentration.
                           535-16

-------
         9.2.5   DL DETERMINATION (optional) - While DL determination is not a
                specific requirement of this method, it may be required by various
                regulatory bodies associated with compliance monitoring. It is the
                responsibility of the laboratory to determine ifDL determination is
                required based upon the intended use of the data.

                Replicate analyses for this procedure should be done over at least 3 days
                (both the sample extraction and the LC/MS/MS analyses should be done
                over at least 3 days).  Prepare at least 7 replicate LFBs using solutions at a
                concentration estimated to be near the DL. This concentration may be
                estimated by selecting a concentration at 2-5 times the noise level.  The
                fortification amount used to obtain the DLs in Table 5 can be used as a
                guide, however the appropriate concentration will depend on the
                sensitivity of the LC/MS/MS system used. Ammonium chloride must be
                added to these samples as described in Section  8.1.1. Analyze the seven
                replicates through all steps of Section 11.  (NOTE:  If an MRL verification
                data set meets these requirements, a DL may be calculated from the MRL
                verification data, and no additional analyses are necessary). Calculate the
                DL using the following equation

                       Detection Limit = St(n. l^ _ alpha = 099)

                where
                       t(n-u-aipha = o.99) = Student's t value for the 99% confidence level
                       with n-1 degrees of freedom
                       n = number of replicates, and
                       S = standard deviation of replicate analyses.

                NOTE: Do not subtract blank values when performing DL calculations.
                The DL is a statistical determination of precision only.2  If the DL
                replicates are fortified at a low enough concentration, it is likely that they
                will not meet the precision and accuracy criteria for CCCs, and may result
                in a calculated DL that is higher than the fortified concentration.
                Therefore, no precision and accuracy criteria are specified.

         9.2.6   CALIBRATION CONFIRMATION - If a QCS is available, analyze the
                QCS as described in Section 9.11 to confirm the accuracy of the
                standards/calibration curve.

9.3      LABORATORY REAGENT BLANK (LRB) - An LRB is required with each
         extraction batch (Sect. 3.6) of samples to determine any background system
         contamination. If within the retention time window of any analyte, the LRB
         produces a peak that would prevent the determination of that analyte, determine
                                    535-17

-------
         the source of contamination and eliminate the interference before processing
         samples. Background contamination must be reduced to an acceptable level
         before proceeding. Background from target analytes or contaminants that
         interfere with the measurement of target analytes must be < 1/3 the MRL. If the
         target analytes are detected in the LRB at concentrations equal to or greater than
         this level, then all data for the problem analyte(s) must be considered invalid for
         all samples in the extraction batch. Any time a new batch of SPE materials is
         received, or new supplies of other reagents are used, repeat the demonstration of
         low system background described in Section 9.2.1.

         NOTE: Although quantitative data below the MRL may not be reliably accurate
         for data reporting, such data is useful in determining an MRL cut off for analytes
         that are typically detected in LRBs. Therefore, blank contaminantion levels may
         be estimated by extrapolation, when the concentration is below the lowest
         calibration standard.

9.4      CONTINUING CALIBRATION  CHECK (CCC) - CCC Standards are analyzed
         at the beginning of each analysis batch, after every ten field samples, and at the
         end of the analysis batch.  See Section 10.3 for concentration requirements,
         frequency requirements, and acceptance criteria.

9.5      LABORATORY FORTIFIED BLANKS (LFB) - With each extraction batch,
         extract and analyze an LFB containing each analyte of concern.  If more than 20
         field samples are included in a batch, analyze an LFB for every 20 samples.  The
         fortified concentration of the LFB must be rotated between low, medium, and
         high concentrations from batch to batch.  The low concentration LFB must be as
         near as practical to the MRL. Similarly, the high concentration LFB should be
         near the high end of the calibration range established during the initial calibration
         (Sect. 10.2). Results of LFB analyses corresponding to the low fortification
         concentration for an analyte must be within 50-150% of the true value for all
         analytes. Results of LFB analyses from medium and high level concentrations
         must be 70-130% of the true value for all analytes.  If the LFB results do not meet
         these criteria for target analytes, then all data for the problem analyte(s) must be
         considered invalid for all samples in the extraction batch.

9.6      INTERNAL STANDARDS (IS) - The analyst must monitor the peak area of the
         IS in all injections during each analysis batch. The IS response (as indicated by
         peak area) in each chromatographic run must not deviate by more than ±50% from
         the average area measured during the initial calibration for the IS.  A poor
         injection, as well as matrix enhancement or suppression, could cause the IS area
         to exceed these criteria. Inject a second aliquot of the suspect extract to determine
         whether the failure is due to poor injection.
                                    535-18

-------
         9.6.1   If the reinjected aliquot produces an acceptable internal standard response,
                report results for that injection.

         9.6.2   If the IS area for the reinjected extract deviates greater than 50% from the
                initial calibration average, the analyst should check the continuing
                calibration check standards that ran before and after the sample. If the
                continuing calibration check fails the criteria of Section 10.3, recalibration
                is in order per Section 10.2, followed by sample reanalysis. If the
                calibration standard is acceptable, extraction of the sample should be
                repeated provided the sample is still within holding time.  If reextraction is
                not possible, sample recollection should be ordered if data is supporting a
                regulatory monitoring requirement. Otherwise, report results obtained
                from the reinjected extract, but annotate as suspect.

9.7      SURROGATE (SUR) RECOVERY - The SUR PDS is fortified into all
         calibration standards, samples, LFBs, LFSMs, LFSMDs, FDs, and LRBs. The
         SUR is also added to the calibration curve and calibration check standards.  The
         SUR is a means of assessing  method performance from extraction to final
         chromatographic measurement.

         9.7.1   Surrogate recovery criteria are 70-130% of the fortified amount for the
                method surrogate. When surrogate recovery from a sample, blank, or
                CCC does not meet these criteria, check : (1) calculations to locate
                possible errors, (2) standard solutions  for degradation, (3) contamination,
                and (4) instrument performance. Correct any problems that are identified.
                If these steps do not reveal the cause of the problem, reanalyze  the extract.

         9.7.2   If the reanalysis meets the surrogate recovery criterion, report only data for
                the reanalyzed extract.

         9.7.3   If the extract reanalysis fails the 70-130% recovery criterion, the analyst
                should check the  calibration by reinjecting the most recently acceptable
                calibration standard.  If the calibration standard fails the criteria of
                Section 10.3, recalibration is in order (Sect. 10.2), followed by sample
                reanalysis.  If the calibration standard  is acceptable and a duplicate sample
                is available, extraction and analysis of the duplicate should be performed
                provided the sample is still within the holding time. If a duplicate sample
                is not available, sample recollection should be ordered if the data is
                supporting a regulatory monitoring requirement.  If this sample reanalysis
                also fails the recovery criterion, report all data for that sample as suspect
                due to surrogate recovery failure or potential sample matrix effect.
                                     535-19

-------
9.8      LABORATORY FORTIFIED SAMPLE MATRIX (LFSM) - Analysis of an
         LFSM is required in each extraction batch and is used to determine that the
         sample matrix does not adversely affect method accuracy. Assessment of method
         precision is accomplished by analysis of a Field Duplicate (Sect. 9.9), however,
         infrequent occurrence of target analytes would hinder this assessment. If the
         occurrence of target analytes in the samples is infrequent, or if historical trends are
         unavailable, a second LFSM, or LFSMD, must be prepared, extracted, and
         analyzed from a duplicate of the field sample. Extraction batches that contain
         LFSMDs will not require the extraction of a Field Duplicate. If a variety of
         different sample matrices are analyzed regularly, for example, drinking water
         from groundwater and surface water sources, method performance should be
         established for each.  Over time, LFSM data should be documented for all routine
         sample sources  for the laboratory.

         9.8.1   Within each extraction batch, a minimum of one field sample is fortified
                as an LFSM for every 20 samples extracted. The LFSM is prepared by
                spiking a sample with an appropriate amount of the analyte PDS
                (Sect. 7.2.3.2). Select a spiking concentration that is greater than or equal
                to the matrix background concentration, if known. Use historical data and
                rotate through the designated concentrations when selecting a fortifying
                concentration, such that fortified concentrations range from near the MRL
                to the high level standard. LFSM recoveries may fall outside the
                acceptance limits (Sect. 9.8.3) if an analyte is present in the sample at a
                concentration that greatly exceeds the spiking concentration.

         9.8.2   Calculate the percent recovery (R) for each analyte using the equation
                                        c


                where
                        A = measured concentration in the fortified sample
                        B = measured concentration in the unfortified sample, and
                        C = fortification concentration.

         9.8.3   Analyte recoveries may exhibit a matrix bias. For samples fortified at or
                above their native concentration, recoveries should be between 60-140%,
                except for low-level fortification near or at the MRL (within a factor of
                two times the MRL concentration), where 50-150% recoveries are
                acceptable. If the accuracy of any analyte falls outside the designated
                range, and the laboratory performance for that analyte is shown to be in
                control in the CCCs, the recovery is judged to be matrix biased. The result


                                    535-20

-------
               for that analyte in the unfortified sample is labeled suspect/matrix to
               inform the data user that the results are suspect due to matrix effects.

9.9      FIELD DUPLICATE OR LABORATORY FORTIFIED SAMPLE MATRIX
         DUPLICATE (FD or LFSMD) - Within each extraction batch, a minimum of one
         Field Duplicate (FD) or Laboratory Fortified Sample Matrix Duplicate (LFSMD)
         must be analyzed. Duplicates check the precision associated with sample
         collection, preservation, storage, and laboratory procedures.  If target analytes are
         not routinely observed in field samples, an LFSMD must be analyzed rather than a
         FD. Refer to Section 9.8.1 for guidance on spiking concentrations.

         9.9.1  Calculate the relative percent difference (RPD) for duplicate
               measurements (FD1 and FD2) using the equation

                                 FD1-FD2
                         RPD =	*100
                                (FDl + FD2)/2

         9.9.2  If an LFSMD is analyzed instead of a Field Duplicate, calculate the
               relative percent difference (RPD) for duplicate LFSMs (LFSM and
               LFSMD) using the equation

                                 ILFSM  - LFSMD
                         RPD = -!	*1QO
                                (LFSM +LFSMD)/2

         9.9.3  RPDs for FDs, or duplicate LFSMs, should fall in the range of 0-30% for
               samples fortified at or above their native concentration.  Greater variability
               may be observed when LFSMs are spiked near the MRL. At the MRL,
               RPDs should fall in the range of 0-50% for samples fortified at or above
               their native concentration. If the RPD of any analyte falls outside the
               designated range, and the laboratory performance for that analyte is shown
               to be in control in the LFB, the recovery is judged to be matrix biased.
               The result for that analyte in the unfortified sample is labeled
               suspect/matrix to inform the data user that the results are suspect due to
               matrix effects.

9.10     RESOLUTION CHECK - The resolution of alachlor ESA and acetochlor ESA, in
         a calibration standard or CCC near or above mid level calibration, must be
         monitored in each analysis batch.  These closely eluting peaks, which are not
         baseline resolved, must have a resolution (Rs) of 1.0 or greater using the equation
                                 3   w   + w
                                     "0.5,1  "0.5,2

                                    535-21

-------
             where
                    tj and t2 = retention times of the first and second adjacent peaks in minutes
                    W0 5 j  and W0 5 2 = widths of the adjacent peaks at half height in minutes.

             At a minimum, resolution must be monitored once for every 24-hour analysis
             batch and may be monitored at any time during the 24-hour period.  It is
             recommended that resolution be checked prior to sample analysis, especially if the
             system in use has a history of resolution problems. If a resolution check fails, the
             problem must be corrected and all samples analyzed since the most recent
             successful calculation of the resolution check, including the QC samples, must be
             reanalyzed.

             Note: Alachlor ESA and acetochlor ESA are structural isomers and have nearly
             identical product ions in MS/MS.  There are dissimilar products ions at m/z 176
             (alachlor ESA) and m/z 162 (acetochlor ESA); however, their relative intensity is
             typically less than 20%. If sufficient sensitivity for these dissimilar ions can be
             achieved, these ions may be used for quantitation and the resolution requirement
             is not necessary.

    9.11     QUALITY CONTROL SAMPLE  (QCS) - The QCS is required only if an
             alternate source is available for these target analytes. Each time that new PDSs
             are prepared, analyze a QCS from  an external source. If standards are prepared
             infrequently,  analyze a QCS at least quarterly.  The QCS may be injected as a
             calibration standard, or fortified into reagent water and analyzed as an LFB.  If the
             QCS is analyzed as a calibration check standard, then the acceptance criteria are
             the same as for the CCC (Sect. 10.3.3).  If the QCS is analyzed as an LFB, then
             the acceptance criteria are the same as for an LFB (Sect. 9.5).  If measured analyte
             concentrations are not of acceptable accuracy, check the entire analytical
             procedure to  locate and correct the problem.

10. CALIBRATION AND STANDARDIZATION

    10.1     After initial calibration is successful, a Continuing Calibration Check (CCC) is
             required at the beginning and end of each analysis batch, and after every tenth
             sample (Sect. 10.3). Initial  calibration should be repeated each time a major
             instrument modification or maintenance is performed.

    10.2     INITIAL CALIBRATION

             10.2.1 ESI-MS/MS TUNE

                    10.2.1.1 Calibrate the mass scale of the MS with the calibration
                            compounds and procedures prescribed by the manufacturer.
                                        535-22

-------
       10.2.1.2 Optimize the [M-H]~ for each target analyte by infusing
              approximately 0.5-1.0 ug/mL of each analyte (prepared in the
              initial mobile phase conditions) directly into the mass
              spectrometer at the LC mobile phase flow rate (approximately
              0.4 mL/min). This tune can be done on a mix of the target
              analytes. The MS parameters (capillary voltage, temperatures,
              gas flows, etc.) are varied until optimal analyte responses are
              determined. The ESAs and OAs may have different optima
              requiring some compromise between the optima. See Table 2 for
              ESI-MS conditions used in method development.

       10.2.1.3 Optimize the product ion (Sect. 3.18) for each target by infusing
              approximately 0.5-1.0 ug/mL of each analyte (prepared in the
              initial mobile phase conditions) directly into the mass
              spectrometer at the LC mobile phase flow rate (approximately
              0.4 mL/min). This tune can be done on a mix of the target
              analytes. The MS/MS parameters (collision gas pressure,
              collision energy, etc.) are varied until optimal analytes responses
              are determined. Typically, all the ESAs have very similar
              MS/MS conditions and all the OAs have similar MS/MS
              conditions. See Tables 3 and 4 for MS/MS conditions used in
              method development.

10.2.2  Establish LC operating parameters equivalent to the suggested conditions
       in Section 17, Table 1.

10.2.3  Inject a mid-level CAL standard under LC/MS conditions to obtain the
       retention times of each target analyte. Ensure that the LC conditions are
       adequate to meet the resolution check in Section 9.10 for alachlor ESA
       and acetochlor ESA. Divide the chromatogram into retention time
       windows each of which contains one or more chromatographic peaks.
       During MS/MS analysis, fragment a small number of selected precursor
       ions   ([M-H]~; Sect. 3.16) for the analytes in that window and choose the
       most abundant product ion. The product ions (also the quantitation ion)
       chosen during method development are in Tables 3 and 4, although these
       will be instrument dependent. For maximum sensitivity, small mass
       windows of ±2 daltons (depending on the instrument) around the product
       ion mass were used. If sufficient sensitivity exists to meet the  MRL, wider
       mass ranges may be used to obtain more confirmation ions.

10.2.4  Inject a mid-level CAL standard under LC/MS/MS conditions  to ensure
       that each target analyte is observed in its MS/MS window and  that there is
       at least 20 scans across the peak for optimum precision. Ensure that there
                           535-23

-------
         are enough scans on either side of the chromatographic peak to account for
         shifts in retention times due to injection imprecision.

  10.2.5  Prepare a set of at least 5 CAL standards as described in Section 7.2.4.
         The lowest concentration CAL standard must be at or below the MRL,
         which may depend on system sensitivity. It is recommended that at least
         four of the CAL standards are at a concentration greater than or equal to
         the MRL.

  10.2.6  The LC/MS/MS system is calibrated using the internal standard technique.
         Concentrations may be calculated through the use of average relative
         response factor (RRF) or through the use of a calibration curve. Calculate
         the RRFs using the equation
         where
                  Ax =  integrated abundance (peak area) of the quantitation ion
                         of the analyte.
                  Ais =   integrated abundance (peak area) of the IS quantitation
                         ion.
                  Qx =   quantity of analyte injected in ng or concentration units.
                  Qis =   quantity of internal standard injected in ng or
                         concentration units.
                  RRF = relative response factor

        Average RRF calibrations may only be used if the RRF values over the
        calibration range are relatively constant. Average RRF is determined by
        calculating the mean RRF of a minimum of five calibration concentrations.
        Average RRF calibrations may only be used if the criteria in Section 10.2.8
        are met.

10.2.7   As an alternative to calculating average RRFs, use the LC/MS/MS data
        system software to generate a linear regression or quadratic calibration
        curve.  Forcing the calibration curve through the origin is not
        recommended. The triple quadrupole instrument used during method
        development yielded linear curves for the target analytes over the
        concentration range of interest.  However, the ion trap instrument yielded
        weighted (1/X) quadratic curves.  Data may be fit with either a linear
        regression (response vs. concentration) or quadratic fit (response vs.
        concentration) and weighting may be used, if necessary.
                              535-24

-------
      10.2.8  When quantitated using the initial calibration curve, each calibration point,
             except the lowest point, for each analyte should calculate to be within
             70-130% of its true value.  The lowest calibration point should calculate to
             be within 50-150% of its true value.  If these criteria cannot be met, the
             analyst will have difficulty meeting ongoing QC criteria. It is
             recommended that corrective action is taken to reanalyze the calibration
             standards, restrict the range of calibration, or select an alternate method of
             calibration.

             CAUTION: When acquiring MS/MS data, LC operating conditions must be
             carefully reproduced for each analysis to provide reproducible retention
             times. If this is not done, the correct ions will not be monitored at the
             appropriate times.  As a precautionary measure, the chromatographic peaks
             in each window must not elute too close to the edge of the time window.

10.3   CONTINUING CALIBRATION CHECK (CCC) - Minimum daily calibration
      verification is as follows. Verify the initial calibration at the beginning and end of
      each group of analyses, and after every tenth sample during analyses. In this context,
      a "sample" is considered to be a field sample.  LRBs, CCCs, LFBs,  LFSMs, FDs and
      LFSMDs are not counted as samples.  The beginning CCC each day must be at or
      below the MRL in order to verify instrument sensitivity prior to any analyses. If
      standards have been prepared such that all low CAL points are not in the same CAL
      solution, it may be necessary to analyze two CAL standards  to meet this
      requirement.  Alternatively, the analyte concentrations in the analyte PDS may be
      customized to meet this criteria. Subsequent CCCs should alternate between a
      medium and high concentration CAL standard.

      10.3.1  Inject an aliquot of the appropriate concentration CAL standard and analyze
             with the same conditions used during the initial calibration.

      10.3.2  Determine that the absolute area of the quantitation ion of the internal
             standard has not changed by more than ± 50% from the average area
             measured during initial calibration. If the IS area has changed by more than
             ±50%, remedial action maybe necessary (Sect.  10.3.4).  If the loss in area
             cannot be recovered, recalibrate according to Section 10.2 and verify
             sensitivity by analyzing a CCC at or below the MRL (Sect  10.3.3).  Control
             charts are useful aids in documenting system sensitivity changes.

      10.3.3  Calculate the concentration of each analyte and surrogate in the CCC. The
             calculated amount for each analyte for medium  and high level CCCs must
             be ± 30%  of the true value.  The calculated amount for the lowest
             calibration point for each analyte must be within ± 50% of the true value. If
             these conditions do not exist, then all data for the problem analyte must be
                                   535-25

-------
                   considered invalid, and remedial action should be taken (Sect. 10.3.4)
                   which may require recalibration. Any field or QC samples that have been
                   analyzed since the last acceptable calibration verification should be
                   reanalyzed after adequate calibration has been restored, with the following
                   exception. If the CCC fails because the calculated concentration is
                   greater than 130% (150% for the low-level CCC) for a particular
                   target analyte, and field sample extracts show no detection for that
                   target analyte, non-detects may be reported without reanalysis.

           10.3.4   REMEDIAL ACTION - Failure to meet CCC QC performance criteria may
                   require remedial action.  Major maintenance such a cleaning the
                   atmospheric pressure ionization source, cleaning the mass analyzer,
                   replacing the LC column, etc., require returning to the initial calibration
                   step (Sect. 10.2).

11.  PROCEDURE

     11.1  This procedure may be performed manually (Sect. 11.4) or in an automated mode
           (Sect. 11.3) using a robotic or automatic sample preparation device. The data
           presented in Tables 5-11 demonstrate data collected by both manual and automated
           extraction.  Extraction and/or elution steps may not be changed or omitted in the
           manual or automated mode.

     11.2  SAMPLE PREPARATION

           11.2.1   Samples are dechlorinated, collected and stored as presented in Section 8.
                   All field and QC samples must contain the dechlorinating agent listed in
                   Section 8.1.1, including the LRB and LFB.  Determine sample volume.  The
                   sample volume may be measured directly in a graduated cylinder to the
                   nearest 10 mL. To minimize the need to use a different graduated cylinder
                   for each sample, an indirect measurement may be done in one of two ways:
                   by marking the level of the sample on the bottle or by weighing the sample
                   and bottle to the nearest 10 g.  After extraction, proceed to Section 11.6 for
                   final volume determination. The LRB and LFB may be prepared by
                   measuring 250 mL of reagent water with a graduated cylinder or filling a
                   250-mL sample bottle to near the top.

           11.2.2   Add an aliquot of the SUR PDS (Sect.  7.2.2.2) to each sample, cap and
                   invert to mix. During method development, a 5-uL aliquot of the 24 ug/mL
                   SUR PDS was added to 250 mL of sample for a final concentration of
                   0.48 ug/L.
                                        535-26

-------
      11.2.3  In addition to the SUR and dechlorination agent, if the sample is an LFB,
             FD, LFSM, or LFSMD, add the necessary amount of analyte PDS
             (Sect. 7.2.3.2). Cap and invert each sample to mix.

      11.2.4  Continue to Section 11.3 if performing automated extractions or 11.4 if
             performing manual extractions.

11.3  AUTOMATED CARTRIDGE SPE PROCEDURE

      11.3.1  The parameters used during method development are shown in the table
             below. Minor modifications to the automated method  are allowed
             providing the QC criteria in Section 9 are met. Modifications to the
             solvents used and the volumes of solvents used is not permitted.
Zvmark AutoTrace Extraction Workstation Procedure
Step 1:
Step 2:
Step 3:
Step 4:
Step 5:
Step 6:
Step 7:
Step 8:
Step 9:
Step 10:
Step 11:
Step 12:
Step 13:
Process 6 samples using the following procedure:
Condition cartridge with 10 mL of 10 mM ammonium
Condition cartridge with 10 mL of 10 mM ammonium
Condition cartridge with 10 mL of reagent water
Condition cartridge with 10 mL of reagent water
Condition cartridge with 10 mL of reagent water
Load 275 mL of sample onto cartridge**
Rinse cartridge with 5 mL of reagent water
Dry cartridge with nitrogen gas for 3 minutes

acetate/MeOH
acetate/MeOH






Soak and collect 5-mL fraction using 10 mM ammonium acetate/MeOH
Collect 10-mL fraction using 10 mM ammonium acetate/MeOH
Clean each sample path with 50 mL of reagent water
End
Flow Rates
Condition: 10 mL/min
Load:
Rinse:
10 mL/min
5 mL/min
Elute: 5 mL/min
Cond Air Push: 10 mL/min
Rinse Air Push: 10 mL/min
Elute Air Push: 5 mL/min


SPE Parameters
Push Delay:
Air Factor:
Autowash Vol.:

5 sec
0.0
2.00 mL

*A sampl
             e volume larger than the actual sample size is entered to ensure that all of the sample is
                                  535-27

-------
      11.3.2  After automated collection of the eluant into the collection tubes, transfer
             the collection tubes to the nitrogen evaporator and proceed to Section 11.5.

11.4  MANUAL CARTRIDGE SPE PROCEDURE

      11.4.1  CARTRIDGE CLEAN-UP AND CONDITIONING - DO NOT allow
             cartridge packing material to go dry during any of the conditioning steps.
             Rinse each cartridge with 20 mL of 10 mM ammonium acetate/methanol
             (Sect. 7.1.4).  Next, rinse each cartridge with 30 mL of reagent water,
             without allowing the water to drop below the top edge of the packing.  If the
             cartridge goes dry during the conditioning phase, the conditioning must be
             started over. Add approximately 3 mL of reagent water to the cartridge,
             attach the transfer tube, turn on the vacuum, and begin adding sample to the
             cartridge.

      11.4.2  SAMPLE EXTRACTON - Adjust the vacuum so that the approximate flow
             rate is 10-15 mL/min.  Do not allow the cartridge to go dry before all the
             sample has passed through. After all of the sample has passed through the
             cartridge, remove the transfer tubing from the top of the cartridge and rinse
             the cartridge with 5 mL of reagent water. Draw air or nitrogen through the
             cartridge for 3 min at high vacuum (10-15 in. Hg).

      11.4.3  CARTRIDGE ELUTION - Lift the extraction manifold top and insert a
             rack with collection tubes into the extraction tank to collect the extracts as
             they are eluted from the cartridge. Elute the analytes from the cartridge
             with 15 mL of 10 mM ammonium acetate/methanol at a low vacuum
             (5 mL/min used to collect method data), such that the solvent exits the
             cartridge in a dropwise fashion.

11.5  EXTRACT CONCENTRATION - Concentrate the extract to dryness under a gentle
      stream of nitrogen in a heated water bath (60-70 °C) to remove all the ammonium
      acetate/methanol. Add the appropriate amount of 5 mM ammonium acetate/reagent
      water (Sect. 7.1.5) and the IS PDS (Sect. 7.2.1.2) to the collection vial to bring the
      volume to 1 mL (10 uL of the 8 ug/mL IS PDS for an extract concentration of
      80 ug/L used for method development). Transfer to autosampler vial.

11.6  SAMPLE VOLUME DETERMINATION - If the level  of the sample was marked
      on the sample bottle, use a graduated cylinder to measure the volume of water
      required to fill the original sample bottle to the mark made prior to extraction.
      Determine to the nearest 10 mL. If using weight to determine volume, weigh the
      empty bottle to the nearest 10 g and determine the sample weight by subtraction of
      the empty bottle from the original weight (Sect. 11.2.1). Assume a sample density of
      1.0 g/mL. In either case, the sample volume will be used in the final calculations of
      the analyte concentration (Sect. 12.2).

                                  535-28

-------
11.7 EXTRACT ANALYSIS

     11.7.1  Establish operating conditions equivalent to those summarized in Tables
             1-4 of Section 17. Instrument conditions and columns should be optimized
             prior to the initiation of the IDC.  Confirm that resolution of alachlor ESA
             and acetochlor ESA peaks meets the requirements of Section 9.10.

     11.7.2  Establish an appropriate retention time window for each analyte. This
             should be based on measurements of actual retention time variation for each
             compound in standard solutions analyzed on the LC over the  course of time.
             Plus or minus three times the standard deviation of the retention time
             obtained for each compound while establishing the initial calibration and
             completing the IDC can be used to calculate a suggested window size;
             however, the experience of the analyst should weigh heavily on the
             determination of the appropriate retention window size.

     11.7.3  Calibrate the system by either the analysis of a calibration curve (Sect. 10.2)
             or by confirming the initial calibration is still valid by analyzing a
             continuing calibration check as described in Section 10.3. If establishing an
             initial calibration for the first time, complete the IDC as described in
             Section 9.2. Confirm that the system meets the resolution check criteria as
             described in Section 9.10 once for every 24-hour analytical batch.

     11.7.4  Begin analyzing field and QC samples at their appropriate frequency by
             injecting the same size aliquots (100 uL was used in method  development),
             under the same conditions used to analyze the initial calibration.

     11.7.5  At the conclusion of data acquisition, use the same software that was used
             in the calibration procedure to identify peaks of interest in predetermined
             retention time windows. Use the data system software to examine the ion
             abundances of the peaks in the chromatogram.  Identify an analyte by
             comparison of its retention time with that of the corresponding analyte peak
             in a reference standard.  Comparison of the mass spectra is not particularly
             useful given the limited ±1-2 dalton mass range around a single product ion
             for each target analyte.

             CAUTION: Alachlor ESA and acetochlor ESA are not baseline
             resolved under the conditions in this method and they have the same
             quantitation ion. Each chromatogram must be manually checked by
             the analyst to verify that the data system has correctly identified the
             peaks in the alachlor ESA/acetochlor ESA retention time window.

     11.7.6  The analyst must not extrapolate beyond the established calibration range.
             If an analyte peak area exceeds the range of the initial calibration curve, the

                                  535-29

-------
                   extract may be diluted with 5 mM ammonium acetate/reagent water
                   (Sect. 7.1.5) with the appropriate amount of internal standard added to
                   match the original concentration. Re-inject the diluted extract. Incorporate
                   the dilution factor into the final concentration calculations. Acceptable
                   surrogate performance (Sect. 9.7) should be determined from the undiluted
                   sample extract.  The resulting data should be documented as  a dilution, with
                   an increased MRL.

12.   DATA ANALYSIS AND CALCULATION

      12.1  Complete chromatographic resolution is not necessary for accurate and precise
           measurements of analyte concentrations using MS/MS with the exception of
           alachlor ESA and acetochlor ESA (Sect. 9.10).  In validating this method,
           concentrations were calculated by measuring the product ions listed in Tables 3 and
           4. Other ions may be selected at the discretion of the analyst.

      12.2  Calculate analyte and surrogate concentrations, using the multipoint calibration
           established in Section 10. Do not use daily calibration verification data to quantitate
           analytes in samples.  Adjust final analyte concentrations to reflect the actual sample
           volume determined in Section 11.6.

      12.3  Prior to reporting the data, the chromatogram should be reviewed for any incorrect
           peak identification (especially in the case of the closely resolved acetochlor ESA and
           alachlor ESA) or poor integration.

      12.4  Calculations must utilize all available digits of precision, but final reported
           concentrations should be rounded to an appropriate number of significant figures
           (one digit of uncertainty), typically two, and not more than three significant figures.

           Note: Some data in Section 17 of this method are reported with more  than two
           significant figures.  This is done to better illustrate the method performance.

13.   METHOD PERFORMANCE

      13.1  PRECISION, ACCURACY, DLs AND LCMRLs - Demonstration data are
           presented using two commercially available tandem mass spectrometers in two
           separate laboratories.  DLs and LCMRLs are presented in Section  17, Table 5.
           Precision and accuracy are presented for four water matrices: reagent water (Table 6
           and 7); chlorinated (finished)  surface water (Table 8); high TOC chlorinated surface
           water (Table 9); and day zero of the holding time in chlorinated ground water
           (Table 10).

      13.2  EVALUATION OF ADDITIONAL ACETANILIDE DEGRADATES - Propachlor
           sulfinyl acetic acid (SAA) and alachlor SAA were evaluated for inclusion in this

                                         535-30

-------
     method. The two compounds were not included due to low and inconsistent
     recoveries on the carbon cartridges.

13.3  ANALYTE STABILITY STUDIES

     13.3.1   AQUEOUS SAMPLES - Chlorinated ground water samples, fortified with
             target analytes at 0.10-0.80 ug/L, were preserved and stored as required in
             Section 8.  The average of seven replicate analyses, conducted on days 0, 7,
             and 14 are presented in Section 17, Table 10.  These data document the
             14-day sample holding time.

     13.3.2   EXTRACTS - Extracts from the  first day of the holding time study
             described above were stored at 4  °C, and analyzed in replicate (N=7) on
             days 0, 7, 14, 21, and 28. The results of these analyses are presented in
             Table 11 and document the 28-day extract holding time.

13.4  PROBLEM COMPOUNDS

     13.4.1   MATRIX ENHANCEMENT - Propachlor OA and fiufenacet OA were
             prone to ESI matrix enhancement in chlorinated surface waters containing
             approximately 2 mg/L or higher TOC.  ESI matrix enhancement of
             propachlor OA (132%) and fiufenacet OA (158%) was observed in typical
             chlorinated surface water using the ion trap instrument (Table 8) due to
             co-eluting TOC. This matrix enhancement was not observed using the
             triple quadrupole instrument due  to the different LC conditions (Table 1)
             used. The TOC eluted earlier in the chromatogram when using the triple
             quadrupole LC conditions instead of the ion trap LC conditions, thus less
             matrix enhancement was observed using the triple quadruple LC conditions.
             Because ion trap instruments have much longer MS/MS scan cycles than
             triple quadrupole instruments, different LC conditions were necessary on
             the ion trap instrument in order to broaden the chromatographic peaks to
             obtain a minimum of 20 scans across the peak. Additional information on
             LC/MS matrix enhancement can be found in the literature.6'7

             This matrix enhancement is the primary reason for not allowing
             modifications to the LC mobile phase constituents. Where the TOC elutes
             in the chromatogram versus the target analytes will determine the amount of
             matrix enhancement as well as which analytes are affected.  Matrix
             enhancement was well characterized during method development, using the
             LC conditions listed in Table 1, to ensure minimal inferences due to TOC.

     13.4.2   SPE BREAKTHROUGH - Without the presence of matrix enhancement to
             obscure  SPE recovery losses, the data using a triple quadrupole instrument
             in Table 9 illustrates the loss in recovery for propachlor OA (33%) and

                                  535-31

-------
                   flufenacet OA (42%) due to excessive TOC (8.6 mg/L) affecting retention
                   on the graphitized carbon solid phase sorbent. The recoveries of
                   propachlor OA (75%) and flufenacet OA (96%) on the ion trap instrument
                   do not appear to be low, but similar to the values in Table 8, the
                   propachlor OA and flufenacet OA recoveries in Table 9 are really matrix
                   enhanced. Thus, the actual SPE sorbent recoveries for propachlor OA and
                   flufenacet OA are probably more similar to those observed for the triple
                   quadrupole data in Table 9.

      13.5  SECOND LABORATORY DEMONSTRATION - The performance of this method
           was demonstrated on the triple quadrupole instrument by a second laboratory and
           included in Section 17, Tables 5-9. The detection limits and LCMRLs observed on
           the triple quadrupole instrument are higher than the ion trap instrument. Different
           LC/MS/MS systems and ESI interfaces respond very differently to the various
           method parameters (e.g., LC mobile phase and modifiers) and the target analytes.
           For example, a second triple quadrupole LC/MS/MS system investigated during the
           method development was more similar in sensitivity to the ion trap instrument.  The
           authors wish to acknowledge the work of Dr. Glynda Smith of Shaw Environmental
           Incorporated for her participation in the second laboratory demonstration.

14.   POLLUTION PREVENTION

      14.1  This method utilizes SPE technology to remove the analytes from water. It requires
           the use of small volumes of solvent and very small quantities of pure analytes,
           thereby minimizing the potential hazards to both the analyst and the environment
           when compared with the use of large volumes of organic solvents in conventional
           liquid-liquid extractions.

      14.2  For information about pollution prevention that may be applicable to laboratory
           operations, consult "Less is Better: Laboratory Chemical Management for Waste
           Reduction" available from the American Chemical Society, on-line at
           http://membership.acs.org/c/ccs/pub_9.htm.

15.   WASTE MANAGEMENT

      15.1  The analytical procedures described in this method generate relatively small amounts
           of waste since only small amounts of reagents and solvents are used.   The matrices
           of concern are finished drinking water.  However, the Agency requires that
           laboratory waste management practices be conducted consistent with all applicable
           rules and regulations, and that laboratories protect the air, water, and land by
           minimizing and controlling all releases from fume hoods and bench operations.
           Also, compliance is required with any sewage discharge permits and regulations,
           particularly the hazardous waste identification rules and land disposal restrictions.
                                        535-32

-------
16.  REFERENCES

1.    "Statistical Protocol for the Determination of the Single-Laboratory Lowest Concentration
     Minimum Reporting Level (LCMRL) and Validation of the Minimum Reporting Level
     (MRL)," available at www.epa.gov/OGWDW/methods/sourcalt.html.

2.    Glaser, J.A., D.L. Foerst, G.D. McKee, S.A. Quave, and W.L. Budde, "Trace Analyses for
     Wastewaters," Environ. Sci. Technol. 1981, 15, 1426-1435.

3.    "OSHA Safety and Health Standards, General Industry," (29CRF1910).  Occupational
     Safety and Health Administration, OSHA 2206, (Revised, July 1, 2001).

4.    "Carcinogens-Working with Carcinogens," Publication No. 77-206, Department of Health,
     Education, and Welfare, Public Health Service, Center for Disease Control, National
     Institute of Occupational Safety and Health, Atlanta, Georgia, August 1977.

5.    "Safety in Academic Chemistry Laboratories," American Chemical Society Publication,
     Committee on Chemical Safety, 6th Edition, available from the ACS Office of Society
     Services by e-mail at OSS@acs.org.

6.    Leenheer, J.A., C.E. Rostad, P.M. Gates, E.T. Furlong, and I. Ferrer, "Molecular
     Resolution and Fragmentation of Fulvic Acid by Electrospray lonization/Multistage
     Tandem Mass Spectrometry." Anal. Chem. 2001, 73, 1461-1471.

7.    Cahill, J.D., E.T. Furlong, M.R. Burkhardt, D. Kolpin, and L.G. Anderson, "Determination
     of Pharmaceutical Compounds in Surface- and Ground-Water Samples by Solid-Phase
     Extraction and High-Performance Liquid  Chromatography Electrospray lonization Mass
     Spectrometry," J. Chromatogr. A 2004. 1041, 171-180.
                                        535-33

-------
17.  TABLES. DIAGRAMS. FLOWCHARTS. AND DEMONSTRATION  DATA*




    TABLE 1. LC METHOD CONDITIONS
Triple Quadrupole Conditions
Time
(min)

initial
7.0
10.0
18.0
20.0
25.0
25.1
40.0
% 5mM
ammonium
acetate
90.0
80.0
75.0
75.0
20.0
20.0
90.0
90.0
% Methanol


10.0
20.0
25.0
25.0
80.0
80.0
10.0
10.0
Agilent Hypersil 2.1 x 100 mm packed
with 5.0 um Clg stationary phase
Flow rate of 0.25 mL/min
Column Temperature of 65°C
100 uL injection
Ion Trap Conditions
Time
(min)

initial
5.0
10.0
35.0
35.1
57.0


% 5mM
ammonium
acetate
95.0
72.0
72.0
20.0
95.0
95.0


% Methanol


5.0
28.0
28.0
80.0
5.0
5.0


Agilent Hypersil 2. 1 x 200 mm packed
with 5.0 um Clg stationary phase
Flow rate of 0.4 mL/min
Column Temperature of 65°C
100 uL injection
TABLE 2.    ESI-MS METHOD CONDITIONS
Triple Quadrupole ESI Conditions
Polarity
Capillary needle voltage
Cone gas flow
Nitrogen desolvation gas
Desolvation gas temp.
Cone voltage
Negative ion
-3kV
50 L/hr
500 L/hr
210°C
40V
Ion Trap ESI Conditions
Polarity
Capillary needle voltage
Nitrogen sheath gas
Nitrogen auxiliary gas
Heated capillary

Negative ion
-3kV
80 (unitless)
20 (unitless)
275°C

 * Instrumentation, when specified, does not constitute endorsement. Brand names are included for illustration only.
                                       535-34

-------
        TABLE 3. TRIPLE QUADRUPOLE MS/MS METHOD CONDITIONS
Segment
#b
1
1
2
2
2
3
3
3
4
4
4
4
2
5
Analyte
Propachlor OA
Flufenacet OA
Propachlor ESA
Flufenacet ESA
Dimethenamid OA
Dimethenamid ESA
Alachlor OA
Acetochlor OA
Alachlor ESA
Metolachlor OA
Acetochlor ESA
Metolachlor ESA
Dimethachlor ESA (SUR)
Butachlor ESA (IS)
Retention Time
(min)
7.33
8.67
10.01
10.81
13.25
14.87 & 15. 11 e
15.86
16.34
18.46
18.60
19.12
20.95
12.18 e
36.95
Precursor
Ion (m/z)
206
224
256
274
270
320
264
264
314
278
314
328
300
356
Product
Ion (m/z)c'd
134
152
80
80
198
80
160f
146f
80
206
80
80
80
80
Collision
Energy
(ev)
8
10
25
25
10
25
10
10
25
8
30
25
25
25
a Argon used as collision gas at 2 mtorr.
b Segments are time durations in which single or multiple scan events occur.
c Ions used for quantitation purposes.
d Scan width was ±0.5 daltons.
  Analyte has two partially resolved chromatographic peaks due to stereoisomers. Both peaks
  summed for quantitation purposes.
f Scan range was 145-161 daltons.
                                        535-35

-------
               TABLE 4. ION TRAP MS/MS METHOD CONDITIONS
Segment
#b
1
2
3
4
5
6
7
7
8
8
8
9
4
10
Analyte
Propachlor OA
Flufenacet OA
Propachlor ESA
Flufenacet ESA
Dimethenamid OA
Dimethenamid ESA
Alachlor OA
Acetochlor OA
Alachlor ESA
Metolachlor OA
Acetochlor ESA
Metolachlor ESA
Dimethachlor ESA (SUR) f
Butachlor ESA (IS)
Retention
Time (min)
11.20
11.95
12.58
13.01
13.43
14.34
15.27
16.46
16.97
18.99
19.24
19.39
13.38
29.26
Precursor
Ion (m/z) c
206
224
256
274
270
320
264
264
314
278
314
328
300
356
Product
Ion (m/z) d'e
134
152
121
121
198
121
160g
146s
121
206
121
121
121
121
Normalized
Collision
Energy (%)
28
28
36
36
28
36
28
28
36
28
36
36
36
36
a An ion trap LC/MS/MS chromatogram of the analytes is shown in Figure 1.
b Segments are time durations in which single or multiple scan events occur.
c Isolation width was ±3 daltons
 Ions used for quantitation purposes.
e Scan width was ±2 daltons.
 Analyte has two partially resolved chromatographic peaks due to stereoisomers. Both peaks
 summed for quantitation purposes.
8 Scan range was 145-161 daltons.
                                       535-36

-------
TABLE 5.  DETECTION LIMITS AND LOWEST CONCENTRATION MRL IN REAGENT WATER

Analyte
Propachlor OA
Flufenacet OA
Propachlor ESA
Flufenacet ESA
Dimethenamid OA
Dimethenamid ESA
Alachlor OA
Acetochlor OA
Alachlor ESA
Metolachlor OA
Acetochlor ESA
Metolachlor ESA
Triple Quadrupole MS a
DL Fortification
Level (|ug/L)
0.20
0.10
0.23
0.21
0.18
0.24
0.20
0.25
0.20
0.20
0.18
0.23
DLcOig/L)
0.093
0.057
0.11
0.099
0.063
0.093
0.12
0.11
0.082
0.090
0.095
0.11
LCMRLd
fog/L)
0.55
0.22
0.40
0.35
0.41
0.45
0.52
0.48
0.32
0.44
0.41
0.40
Ion Trap MS b
DL Fortification
Level (jig/L)
0.080
0.010
0.020
0.020
0.020
0.040
0.040
0.040
0.020
0.010
0.020
0.080
DLc(ug/L)
0.036
0.012
0.010
0.0060
0.013
0.029
0.032
0.019
0.0080
0.010
0.012
0.043
LCMRLd
fog/L)
0.10
0.016
0.035
0.034
0.031
0.048
0.076
0.074
0.032
0.017
0.030
0.11
a Samples extracted manually and analyzed using conditions in Table 3.
 Samples extracted by automation and analyzed using conditions in Table 4.
c Detection Limits were determined by analyzing 7 replicates over 3 days using the conditions outlined in Tables 1-4.
d LCMRLs were calculated according to the procedure in reference 1.
                                                     535-37

-------
          TABLE 6. PRECISION AND ACCURACY OF LOW LEVEL FORTIFIED REAGENT WATER

Analyte
Propachlor OA
Flufenacet OA
Propachlor ESA
Flufenacet ESA
Dimethenamid OA
Dimethenamid ESA
Alachlor OA
Acetochlor OA
Alachlor ESA
Metolachlor OA
Acetochlor ESA
Metolachlor ESA
Dimethachlor (SUR)
Triple Quadrupole MS a
Fortification
Level (jig/L)
0.50
0.26
0.59
0.54
0.45
0.59
0.50
0.62
0.51
0.49
0.44
0.58
1.8
Mean %
Recovery
89
88
89
88
74
78
76
82
83
77
84
87
94
RSD (%)
7.5
6.3
11
9.3
12
15
3.9
4.7
7.2
6.0
5.8
7.0
7.8
Ion Trap MS b
Fortification
Level (jig/L)
0.16
0.020
0.040
0.040
0.040
0.080
0.080
0.080
0.040
0.020
0.040
0.16
0.48
Mean %
Recovery
98
103
103
108
98
98
93
99
108
91
103
101
108
RSD (%)
3.9
15
14
14
16
7.7
17
15
13
16
13
10
7.6
1 Samples (N=7) extracted manually and analyzed using conditions in Table 3.
' Samples (N=7) extracted by automation and analyzed using conditions in Table 4.
                                                  535-38

-------
         TABLE 7. PRECISION AND ACCURACY OF HIGH LEVEL FORTIFIED REAGENT WATER

Analyte
Propachlor OA
Flufenacet OA
Propachlor ESA
Flufenacet ESA
Dimethenamid OA
Dimethenamid ESA
Alachlor OA
Acetochlor OA
Alachlor ESA
Metolachlor OA
Acetochlor ESA
Metolachlor ESA
Dimethachlor (SUR)
Triple Quadrupole MS a
Fortification
Level (|ug/L)
2.0
1.0
2.3
2.1
1.8
2.4
2.0
2.5
2.0
2.0
1.7
2.3
1.8
Mean %
Recovery
89
90
100
98
85
94
80
82
92
85
96
94
95
RSD (%)
5.8
4.2
1.8
1.4
3.3
3.1
6.2
4.6
2.4
6.5
3.3
2.8
1.6
Ion Trap MS b
Fortification
Level (jig/L)
0.80
0.10
0.20
0.20
0.20
0.40
0.40
0.40
0.20
0.10
0.20
0.80
0.48
Mean %
Recovery
102
106
106
104
96
108
80
88
100
91
108
104
100
RSD (%)
3.6
3.2
3.9
3.5
7.1
6.0
9.9
9.9
5.3
12
7.7
9.4
5.6
1 Samples (N=7) extracted manually and analyzed using conditions in Table 3.
' Samples (N=6) extracted by automation and analyzed using conditions in Table 4.
                                                  535-39

-------
TABLE 8. PRECISION AND ACCURACY IN CHLORINATED SURFACE WATER

Analyte
Propachlor OA
Flufenacet OA
Propachlor ESA
Flufenacet ESA
Dimethenamid OA
Dimethenamid ESA
Alachlor OA
Acetochlor OA
Alachlor ESA
Metolachlor OA
Acetochlor ESA
Metolachlor ESA
Dimethachlor (SUR)
Triple Quadrupole MS b
Fortification
Level (|ug/L)
1.5
0.78
1.8
1.6
1.4
1.8
1.5
1.9
1.5
1.5
1.3
1.7
1.8
Mean %
Recovery
63
68
81
85
73
82
84
83
80
69
80
72
81
RSD (%)
1.4
2.0
2.4
1.7
1.4
2.4
1.8
1.8
2.0
2.7
1.9
3.8
1.2
Ion Trap MS c
Fortification
Level (jig/L)
0.80
0.10
0.20
0.20
0.20
0.40
0.40
0.40
0.20
0.10
0.20
0.80
0.48
Mean %
Recovery
132
158
93
103
99
106
97
103
119
89
117
106
100
RSD (%)
12
14
6.0
1.8
8.8
3.3
11
10
4.3
4.7
2.9
2.0
5.6
a TOC = 1.6 mg/L and hardness = 120 mg/L.
b Samples (N=6) extracted manually and analyzed using conditions in Table 3.
c Samples (N=4) extracted by automation and analyzed using conditions in Table 4.
                                                     535-40

-------
TABLE 9. PRECISION AND ACCURACY IN HIGH TOC CHLORINATED SURFACE WATER

Analyte
Propachlor OA
Flufenacet OA
Propachlor ESA
Flufenacet ESA
Dimethenamid OA
Dimethenamid ESA
Alachlor OA
Acetochlor OA
Alachlor ESA
Metolachlor OA
Acetochlor ESA
Metolachlor ESA
Dimethachlor (SUR)
Triple Quadrupole MS b
Fortification
Level (|ug/L)
1.5
0.78
1.8
1.6
1.4
1.8
1.5
1.9
1.5
1.5
1.3
1.7
1.8
Mean %
Recovery
33
42
87
91
60
82
63
64
87
67
89
77
83
RSD (%)
4.0
4.5
2.4
2.7
2.4
3.0
3.0
3.1
2.9
3.8
2.8
3.0
2.4
Ion Trap MS c
Fortification
Level (jig/L)
0.80
0.10
0.20
0.20
0.20
0.40
0.40
0.40
0.20
0.10
0.20
0.80
0.48
Mean %
Recovery
75
96
71
84
83
73
88
73
74
122
84
109
86
RSD (%)
11.6
10.1
6.5
8.8
6.1
4.2
4.5
8.6
9.4
2.5
6.6
5.6
4.1
a TOC = 8.6 mg/L.
b Samples (N=7) extracted manually and analyzed using conditions in Table 3.
c Samples (N=4) extracted by automation and analyzed using conditions in Table 4.
                                                     535-41

-------
     TABLE 10. SAMPLE HOLDING TIME DATA (N=7) FOR CHLORINATED GROUND WATER SAMPLES
                                                                                                     a,b
Analyte
Propachlor OA
Flufenacet OA
Propachlor ESA
Flufenacet ESA
Dimethenamid OA
Dimethenamid ESA
Alachlor OA
Acetochlor OA
Alachlor ESA
Metolachlor OA
Acetochlor ESA
Metolachlor ESA
Fortification
Level (|ug/L)
0.80
0.10
0.20
0.20
0.20
0.40
0.40
0.40
0.20
0.10
0.20
0.80
DayO
Mean %
Recovery
95
116
106
101
105
105
94
100
89
91
91
97
RSD (%)
3.7
2.4
3.6
4.9
1.8
1.3
5.3
4.1
4.9
4.1
5.0
4.9
Day 7
Mean %
Recovery
112
128
112
112
107
108
102
103
105
96
104
104
%RSD
6.3
4.5
5.0
6.2
6.0
4.3
5.9
4.8
1.9
5.6
6.9
4.5
Day 14
Mean %
Recovery
107
128
102
107
108
100
100
96
93
90
82
97
RSD (%)
10
8.2
9.4
7.9
4.9
5.8
10
8.3
3.3
4.5
5.9
4.7
a TOC = 1.6 mg/L and hardness =360 mg/L.
b Samples dechlorinated, extracted by automation and analyzed using ion trap conditions in Table 4.
                                                    535-42

-------
TABLE 11. EXTRACT HOLDING TIME DATA (N=7) FOR CHLORINATED GROUND WATER SAMPLES
                                                                                                  a,b
Analyte
Propachlor OA
Flufenacet OA
Propachlor ESA
Flufenacet ESA
Dimethenamid OA
Dimethenamid ESA
Alachlor OA
Acetochlor OA
Alachlor ESA
Metolachlor OA
Acetochlor ESA
Metolachlor ESA
Dimethachlor (SUR)
Fortification
Level (|ug/L)
0.80
0.10
0.20
0.20
0.20
0.40
0.40
0.40
0.20
0.10
0.20
0.80
0.48
DayO
Mean
%Rec
95
116
106
101
105
105
94
100
89
91
91
97
102
RSD
(%)
3.7
2.4
3.6
4.9
1.8
1.3
5.3
4.1
4.9
4.1
5.0
4.9
3.9
Day 7
Mean
%Rec
98
112
93
102
98
102
86
86
100
81
92
102
94
RSD
(%)
4.7
4.6
4.5
6.6
3.7
6.3
7.3
11.1
6.4
5.7
6.1
4.9
5.8
Day 14
Mean
%Rec
99
121
95°
107
111
102
101
112
105
103
97
107
103
RSD
(%)
5.5
3.9
4.3
4.2
1.8
3.4
7.2
5.7
5.1
2.7
5.5
4.6
6.7
Day 21
Mean
%Rec
100
114
92C
97
103
97
100
97
94
92
93
98
101
RSD
(%)
5.0
3.0
7.7
7.4
4.6
7.6
9.8
8.9
6.4
4.0
3.9
3.1
6.3
Day 28
Mean
%Rec
95
112
100
100
106
95
104
102
106
101
88
98
103
RSD
(%)
6.1
4.4
6.0
5.5
6.6
4.4
11.4
8.8
4.3
7.6
4.7
5.3
6.1
a TOC = 1.6 mg/L and hardness =360 mg/L.
b Samples dechlorinated, extracted by automation and analyzed using ion trap conditions in Table
CN=6
                                                                           4.
                                                535-43

-------
    TABLE 12.  INITIAL DEMONSTRATION OF CAPABILITY (IDC) REQUIREMENTS
Method
Reference
Requirement
Specification and Frequency
Acceptance Criteria
Sect. 9.2.1
Initial Demonstration of
Low System Background
Analyze LRB prior to any other IDC steps, or
anytime a new lot of SPE materials or
reagents are used.
Demonstrate that all target analytes are

-------
    TABLE 13.  QUALITY CONTROL REQUIREMENTS (SUMMARY)
Method
Reference
Requirement
Specification and Frequency
Acceptance Criteria
Sect. 8.1
Sect 8.4
Sample Collection,
Preservation and
Holding Time
14 days, protected from light, with addition
of ammonium chloride.
Iced or refrigerated at 10 °C or less for up to 48 hours
to allow time for shipping; refrigerated at 6 °C or less
after arrival at the laboratory.
Sect. 8.4
Extract Holding
28 days
Stored at 4 °C or less in amber vials.
Sect. 9.3
Laboratory Reagent
Blank (LRB)
Include a LRB with each extraction batch
(up to 20 samples). Analyze prior to
analyzing samples and determine to be free
from interferences.
Demonstrate that all target analytes are below 1/3 the
MRL, and that possible interference from reagents and
glassware do not prevent the identification and
quantitation of target analytes.
Sect. 9.4 and
Sect. 10.3
Continuing
Calibration Check
(CCC)
Verify initial calibration by analyzing a low
level (at the MRL or below) CCC prior to
analyzing samples. CCCs are then injected
after every 10 samples and after the last
sample, rotating concentrations to cover the
calibrated range of the instrument.
Recovery for each analyte must be within 70-130% of
the true value for all but the lowest level of calibration.
The lowest calibration level CCC must be within
50-150% of the true value.
Sect. 9.5
Laboratory Fortified
Blank (LFB)
Analyze at least one LFB daily or one for
each extraction batch of up to 20 field
samples.  Rotate the fortified concentration
between low, medium and high amounts.
Results of LFB analyses must be 70-130% of the true
value for each analyte and surrogate for all fortified
concentrations greater than the lowest CAL point.
Results of LFBs corresponding to the lowest CAL point
must be 50-150% of the true value.
Sect. 9.6
Internal Standard (IS)
Butachlor ESA is added to all standards
and extracts.
Peak area counts of the IS in LFBs, LRBs and sample
extracts must be within 50-150% of the average peak
area in the initial calibration.
                                                            535-45

-------
Method
Reference
Requirement
Specification and Frequency
Acceptance Criteria
Sect. 9.7       Surrogate Standards
               (SUR)
                       The surrogate, dimethachlor ESA, is added
                       to all calibration standards, samples, LFBs,
                       LFSMs, LFSMDs, FDs, and LRBs.
                                          Surrogate recovery must be within 70-130% of the true
                                          value.
                                          Samples that fail criteria, must be reported as suspect
                                          due to surrogate recovery or potential matrix effect.
Sect. 9.8       Laboratory Fortified
               Sample Matrix
               (LFSM)
                       Analyze one LFSM per extraction batch
                       (20 field samples or less) fortified with
                       method analytes at a concentration > the
                       native concentration.
                                          Recoveries not within 60-140% (50-150% at the MRL)
                                          of the fortified amount may indicate a matrix effect.
Sect. 9.9       Field Duplicates (FD)
               or Laboratory
               Fortified Sample
               Matrix Duplicates
               (LFSMD)
                       Analyze 1 FD for each 20 samples, or 1 per
                       extraction batch, whichever is greater. See
                       the referenced section for a discussion of
                       when LFSMDs should be analyzed.
                                          RPDs should be within 0-30%.
Sect. 9.10
Resolution Check
Monitor once for every 24 hour analysis
period.
Alachlor ESA and acetochlor ESA that are not baseline
resolved must have a resolution of 1.0 or greater using
the equation in Sect. 9.10.
Sect. 9.11      Quality Control
               Sample (if available)
                       Analyze at least quarterly or when
                       preparing new standards, as well as during
                       the IDC.
                                          If analyzed as a calibration sample, CCC criteria apply.
                                          If analyzed as an LFB, those criteria apply.
Sect. 10.2
Initial Calibration
Use internal standard calibration technique
to generate an average RF, or first or
second order calibration curve for each
analyte. Use a minimum of 5 CAL
standards.
When each calibration standard is calculated using the
calibration curve, the results should be 70-130% of the
true value for all but the lowest standard. The lowest
standard should be 50-150% of the true value.
                                                             535-46

-------
   iOOQOQQ -
   4000000 -
   3000000 -
   2000000 -
   1000000 -
       0 -
   1200000
   1000000 •
    800000-

    600000-
    400000-
    200000-
                     11
                                                  SUR
                                  12
                                               13
                                                            14
                                                                                     •~i
                                                                                     16
                                     12
IS
                         18
                                             22
                                           Time (min)
                                                        24
                                                                  26
                                                                            28
                                                                                      30
Target Peaks
1 propachlor OA
2 flufenacet OA
3 prop achlor E SA
4 flufenacet E SA
5 dimethenamid OA
6 dimethenamid ESA
TalachlnrOA
8 acetochlor OA
9alachlorESA
10 metolachbr OA
11 acetochlor ESA
12 metolachbr ESA
Figure 1.   Ion trap LC/MS/MS chromatogram of a standard containing the target analytes at the concentrations listed in
           Table 10. Note there some of the analytes (SUR, 5, and 6) have two partially resolved peaks due to
           steroisomers.
                                                              535-47

-------
                             METHOD 535 APPENDIX A

A.1   SCOPE AND APPLICATION

      This appendix is intended to be used to synthesize butachlor ESA, which is the suggested
      internal standard (IS) used in Method 535. During method development, the following
      directions, adapted from Feng," were used to synthesis the IS; however, these directions
      are not intended to be restrictive.  The following steps maybe modified provided the
      MS/MS mass spectrum matches that provided in Figure Al and that no contaminant peaks
      are present which could interfere in the quantitation of the target analytes.

A.2   SAFETY

      A.2.1  The toxicity or carcinogenicity of each reagent used in this appendix has not been
            precisely defined; each chemical reagent should be treated as a potential health
            hazard, and exposure to these chemicals should be minimized. The laboratory is
            responsible for maintaining a current awareness file of OSHA regulations
            regarding the safe handling of the chemicals specified in this appendix. A
            reference file of MSDSs should also be made available to all personnel involved in
            the chemical analysis.

      A.2.2  Pure standard materials and stock standards should be handled with suitable
            protection to skin and eyes. Care should be taken not to breathe the vapors or
            ingest the materials.

      A.2.3  At a minimum, the refiuxing, acidification and separatory funnel extractions
            should be performed in a chemical hood.

A.3   EQUIPMENT AND SUPPLIES (Brand names and/or catalog numbers  are included for
      illustration only, and do not imply endorsement of the product.)

      A.3.1  GLASSWARE - All glassware must be meticulously cleaned.  Wash glassware
            with detergent and tap water, rinse with tap water, followed by reagent water.  A
            final rinse with solvents may be needed. In place of a solvent rinse,
            non-volumetric glassware can be heated in a muffle furnace at 400 °C for 2 hours.
            Volumetric glassware should not be heated above 120 °C.

      A.3.2  GRADUATED CYLINDERS - Various sizes.

      A.3.3  BALANCE - Analytical, capable of accurately weighing to 0.0001 g.

      A.3.4  DISPOSABLE PIPETTES - Various sizes, used to transfer solutions.
        Feng, P.C.C. (1991) Pestic.  Biochem. Physiol. 40, 136-142.

                                        535-48

-------
     A.3.5  DRYING OF THE PRODUCT

            A.3.5. 1 VACUUM DESICCATOR - To dry the synthesized product (Fisher
                   Cat.# 08-632 or equivalent).

            A.3.5.2 FREEZE DRYER (OPTIONAL) - To dry the synthesized product
                   (Labconco, Freeze Dry System/Freezone 4.5, Cat. # 77510-OON or
                   equivalent).

     A.3.6  ROUND BOTTOM OR FLAT BOTTOM FLASK - 50-mL pyrex flask (or
            equivalent) to contain and reflux the reactants.

     A.3.7  FRIEDRICH CONDENSER - 250 mL, used to prevent evaporation losses during
            refluxing (Kimble-Kontes Cat.# 456250-0022 or equivalent).

     A.3 . 8  SEPARATORY FUNNEL - 60 mL, used in the liquid-liquid extraction of the
            synthesized IS after refluxing.

     A.3. 9  HEATING MANTLE - Sized to fit and heat a 50-mL flask to 85-95 °C.

     A.3. 10 EXTRACT CONCENTRATION SYSTEM - Synthesized product is concentrated
            by blowdown with nitrogen using a water bath set at 60-65 °C (Meyer N-EVAP,
            model 111, Organomation Associates, Inc. or equivalent).

A.4  REAGENTS AND GASES

     A.4. 1  BUTACHLOR (C17H26C1NO2, CAS#: 23 1 84-66-9) - Reactant in IS synthesis
            (Supelco, Cat. #PS-348 or equivalent).

     A.4.2  ETHANOL (C2H6O, CAS#: 64-17-5) - >190 proof (Sigma Cat.# E7148 or
            equivalent).

     A.4.3  BOILING CHIP S (CAS#: 1344-28-1) - Used in refluxing step (Fisher Cat.# B365,
            or equivalent). A stir bar and stir plate can be used in place of the boiling chips, if
            preferred.
     A.4.4  SODIUM SULFITE (Na^Og, CAS#: 7757-83-7) - Reactant in IS synthesis (Fisher
            Cat.# S-447 or equivalent).

     A.4.5  SULFURIC ACID (H2SO4, CAS#:7664-93-9) - Used to acidify synthesized IS
            before extracting (Fisher Cat.# A300-500 or equivalent).

     A.4.6  METHYLENE CHLORIDE (CH2C12, CAS#: 75-09-2) - Extraction solvent (Fisher
            Cat.# D151 or equivalent).
                                      535-49

-------
      A.4.7  NITROGEN - Used to evaporate the methylene chloride (Ultra High Purity or
             equivalent).

      A.4.8  REAGENT WATER - Purified water which does not contain any measurable
             quantities of any target analytes.

A.5   SYNTHESIS OF BUTACHLOR ESA

      A.5.1  Add 13.5 mL of reagent water to a 50-mL round-bottom flask with a few boiling
             chips.

      A.5.2  Add 1.5 mL of ethanol to the flask.

      A.5.3  Add 0.5g of butachlor and 3g of sodium sulfite to the flask containing the ethanol
             and water. Not all the butachlor and sodium sulfite will be soluble and the solution
             may look cloudy.

      A.5.4  Gently heat and reflux the solution overnight (-20 hours).  The solution will
             become clear during the refluxing.

      A.5.5  After refluxing, allow the solution  to cool. Add 2 mL of concentrated sulfuric acid
             to the round-bottom flask containing the refluxed and cooled solution.  At this
             point the solution will become milky white in appearance.

      A.5.6  Transfer the reaction solution to a 60-mL separatory funnel with a disposable
             pipette.  Wash the round-bottom flask with several milliliters of reagent water and
             transfer the washes to the separatory funnel.

      A.5.7  Add 20 mL of methylene chloride to the separatory funnel. Shake the funnel for
             approximately 2 minutes with frequent venting. A milky white emulsion will
             form.  Allow the phases to separate for approximately 24 hours. The emulsion
             should separate into 2 phases: a methylene chloride phase and a water phase.  If the
             emulsion has not cleared, allow the phases to stand longer until separation is
             complete.

      A.5.8  Obtain the dry weight of the collection apparatus (beaker, tube, etc.) and collect the
             bottom layer (methylene chloride layer) of the separatory funnel. Discard the water
             fraction.

      A.5.9  Evaporate the methylene chloride to dryness in a nitrogen evaporator at room
             temperature. After evaporation the residue will appear viscous.  Weigh the
             residue.
                                         535-50

-------
      A.5.10 Place the residue in a vacuum desiccator. Dry the product until a constant weight
             is achieved. This may take several weeks. Other methods of evaporation (such as
             freeze drying) are also acceptable.

A.6   VERIFICATION OF BUTACHLOR ESA

      A.6.1  Follow the procedure in Sect. 7.2.1 of Method 535 to prepare an IS PDS solution
             of the IS near a concentration of 10 ug/mL in methanol.

      A.6.2  Using the IS PDS, prepare a "calibration standard" containing only the synthesized
             IS at an appropriate concentration.  During method development, the IS
             concentration in standards and extracts was 80 ug/L.

      A.6.3  Tune the LC/MS using the procedures outline in Section 10.2.1.1 and 10.2.1.2 of
             Method 535.

      A.6.4  Initially, perform a full scan LC/MS run to locate the IS peak and verify the
             molecular weight (butachlor ESA MW=357). Set the MS to scan at least
             200-400 daltons in the negative ion ESI mode.

      A.6.5  Inject an aliquot of the "calibration standard" into the LC/MS using the suggested
             LC parameters outlined in Section 17, Table  1 of Method 535.  Other LC
             conditions may be used at the discretion of the analyst. If other LC conditions are
             used, more emphasis will have to be placed on the interpretation of the mass
             spectra since the IS retention time will not match the retention time demonstrated
             in Method 535.

      A.6.6  Locate the chromatographic peak containing  the [M-H]~ ion at m/z 356. Note the
             retention time. In addition, carefully observe the whole chromatogram and verify
             that no other peaks (contaminants)  are present which will interfere in the analysis
             of the other target analytes.  Since the final analysis is MS/MS, the main concern at
             this point should be whether any potential contaminants are present in large enough
             quantities to cause suppression in the electrospray source.  A suggested guideline
             is: if the contaminant is present at lOx the IS  level, it may cause suppression of a
             co-eluting  target analyte. If such a contaminant is present, further purification of
             the synthesized IS may be necessary. Figure  Al-A and Al-B demonstrate the
             relative purity of the synthesized standard. A full scan mass spectra of synthesized
             butachlor ESA is demonstrated in Figure Al-C.

      A.6.7  Optimize the LC/MS/MS using the procedures  outlined in Section 10.2.1.3 of
             Method 535  Confirm the identity of the synthesized IS with LC/MS/MS by
             scanning all products ions produced by fragmentation of m/z 356.  Suggested
             MS/MS parameters are found in Tables 3 or 4 of Method 535.  Obtain a full scan
             MS/MS by scanning the product ions over a mass range of 75-357 daltons, if
             possible. (Ion trap instruments may not be able to scan this large of a mass range.

                                         535-51

-------
In that case, use the lowest starting mass the ion trap will allow.) A product ion
mass spectra of synthesized butachlor ESA is demonstrated in Figure A2 using an
ion trap instrument. MS/MS fragmentation and relative abundances can vary
significantly between instruments and depends heavily on various instrument
settings. At a minimum, m/z 121 must be present in the synthesized IS and if a
triple quadrupole instrument is used, m/z 80 must be present. Other ions present in
Figure A2 should be used, at the discretion of the analyst, to further verify the
synthesis of the IS.
                            535-52

-------
                 A) Full Son LC MS
                                                         25.13
                                                         JL
                                                29.34
                                                _ K
                                                     31 .55
   
-------
         1DD -3
          90 -
          70 -=
          60 -
Js
cc. 40
          30 -
          20 -
          10 -
           0 -
                      121
                                             176
                                     158
                                  149
                        122
                        I.
                                       160
                                                                                                                  356
                                                                                                                      357
                                                              218
                                                                                      282 287
              100
                      120
                              140
                                      160
                                              180
                                                      200
                                                              220
                                                                      240
                                                                              260
                                                                                       280
                                                                                               300
                                                                                                       320
                                                                                                               340
                                                                                                                       360
Figure A2.    Product ion mass spectrum of m/z 356 used for confirmation of synthesized butachlor ESA using an ion trap mass
              spectrometer.
                                                           535-54

-------
                               Method 535 APPENDIX B

B.I   SCOPE AND APPLICATION

      This appendix is intended to be used to synthesize dimethachlor ESA which is the
      suggested surrogate (SUR) standard used in Method 535. During method development,
      the following directions, adapted from Feng,a were used to synthesis the SUR; however,
      these directions are not intended to be restrictive.  The following steps may be modified
      provided the MS/MS mass spectrum matches that provided below and that no contaminant
      peaks are present which could interfere in the quantitation of the target analytes.

B.2   SAFETY

      B.2.1 The toxicity or carcinogenicity  of each reagent used in this appendix has not been
           precisely defined; each chemical reagent should be treated as a potential health
           hazard, and exposure to these chemicals should be minimized. The laboratory is
           responsible for maintaining a current awareness file of OSHA regulations regarding
           the safe handling of the chemicals specified in this appendix.  A reference file of
           MSDSs should also be made available to all personnel involved in the chemical
           analysis.

      B.2.2 Pure standard materials and stock standards of these compounds should be handled
           with suitable protection to skin and eyes.  Care should be taken not to breathe the
           vapors or ingest the materials.

      B.2.3 At a minimum, the refluxing, acidification and separatory funnel extractions should
           be performed in a chemical hood.

B.3   EQUIPMENT AND SUPPLIES (Brand names and/or catalog numbers are included for
      illustration only, and do not imply endorsement of the product.)

      B.3.1 GLASSWARE - All glassware must be meticulously cleaned. Wash glassware
           with detergent and tap water, rinse with tap water, followed by reagent  water.  A
           final rinse with solvents maybe needed. In place of a solvent rinse, non-volumetric
           glassware can be heated in a muffle furnace at 400 °C for 2 hours.  Volumetric
           glassware should not be heated above 120 °C.

      B.3.2 GRADUATED CYLINDERS - Various sizes.

      B.3.3 BALANCE - Analytical, capable of accurately weighing to 0.0001 g.
        Feng, P.C.C. (1991) Pestic.  Biochem. Physiol 40, 136-142.

                                        535-55

-------
     B.3.4 DISPOSABLE PIPETTES - Various sizes, used to transfer solutions.

     B.3.5 CONICAL COLLECTION TUBES -15 mL and 50 mL, or other glassware suitable
           for collection of the eluent from the solid phase cartridge after extraction of the
           synthesized SUR.

     B.3.6 SOLID PHASE EXTRACTION (SPE) APPARATUS USING CARTRIDGES

           B.3.6.1  SPE CARTRIDGES - 5.0 g, 20 mL SPE cartridges containing a
                   nonporous graphitized carbon sorbent phase (Supelclean ENVI-Carb SPE
                   cartridges, Supelco Catalog No. 57129 or equivalent).

           B.3.6.2  VACUUM EXTRACTION MANIFOLD - A manual vacuum manifold
                   (Supelco Cat. No. 57030 and 57275 or equivalent) for cartridge
                   extractions.

           B.3.6.3  LABORATORY OR ASPIRATOR VACUUM SYSTEM - Sufficient
                   capacity to maintain a vacuum of approximately 25 cm (10 in) of mercury.

     B.3.7 DRYING OF THE PRODUCT

        B.3.7.1 VACUUM DESICCATOR - To dry the synthesized product (Fisher
              Cat.# 08-632 or equivalent).

     B.3.8 ROUND BOTTOM OR FLAT BOTTOM FLASK - 50-mL pyrex flask (or
           equivalent) to contain and reflux the reactants.

     B.3.9 FRIEDRICH CONDENSER - 250 mL, used to prevent evaporation losses during
           refluxing (Kimble-Kontes Cat.# 456250-0022 or equivalent).

     B.3.10   HEATING MANTLE - Sized to fit and heat a 50-mL flask to 85-95 °C.

     B.3.11   EXTRACT CONCENTRATION SYSTEM - Extracts are concentrated by
              blowdown with nitrogen using a water bath set at 60-65 °C (Meyer N-EVAP,
              model 111, Organomation Associates, Inc. or equivalent).

B.4  REAGENTS AND GASES

     B.4.1 DIMETHACHLOR (C13H18C1NO2,  CAS#: 50563-36-5) - Reactant  in SUR
           synthesis (Sigma-Aldrich, Cat. #45447 or equivalent).

     B.4.2 ETHANOL (C2H6O, CAS#: 64-17-5) - >190 proof (Sigma Cat.#E7148 or
           equivalent).
                                     535-56

-------
      B.4.3 METHANOL (CH3OH, CAS#: 67-56-1) - High purity, demonstrated to be free of
           target analytes and interferences (Fisher Optima grade or equivalent).

      B.4.4 AMMONIUM ACETATE (NH4C2H3O2, CAS#: 631-61-8) -High purity,
           demonstrated to be free of analytes and interferences (Sigma- Aldrich ACS grade or
           equivalent).

      B.4.5 10 mM AMMONIUM ACETATE/METHANOL - To prepare 1L, add 0.7708 g
           ammonium acetate to 1L of methanol. During method development, this solution
           was shown to be stable for at least 18 days.

      B.4.6 BOILING CHIP S (CAS#: 1344-28-1) - Used in refiuxing step (Fisher Cat.# B365,
           or equivalent). A stir bar and stir plate can be used in place of the boiling chips, if
           preferred.
      B.4.7  SODIUM SULFITE (NajSOg, CAS#: 7757-83-7) - Reactant in SUR synthesis
            (Fisher Cat.# S-447 or equivalent).

      B.4.8  NITROGEN - Used to evaporate the methylene chloride (Ultra High Purity or
            equivalent).

      B.4.9  REAGENT WATER - Purified water which does not contain any measurable
            quantities of any target analytes.

B.5   SYNTHESIS OF DIMETHACHLOR ESA

      B.5.1  Add 13.5 mL of reagent water to a 50-mL round-bottom flask with a few boiling
            chips.

      B.5. 2  Add 1.5 mL of ethanol to the flask.

      B.5. 3  Add 0.25 g of dimethachlor and 1.5 g of sodium sulfite to the flask containing the
            ethanol and water.

      B.5. 4  Gently heat and reflux the solution overnight (-20 hours).

      B.5. 5  Allow the refluxed solution to cool.  Divide the 15-mL refiuxed solution into two
            approximately equal portions. Prepare two SPE cartridges according to the
            procedure in Section B. 5.6.1 and extract each portion on the cartridges according to
            the procedure in Section B5.6.2.

      B.5.6  MANUAL CARTRIDGE SPE PROCEDURE
                                        535-57

-------
            B.5.6.1  CARTRIDGE CLEAN-UP AND CONDITIONING - DO NOT allow
                    cartridge packing material to go dry during any of the conditioning steps.
                    Rinse each cartridge with 60 mL of 10 mM ammonium acetate/methanol
                    (Sect. B.4.5). Next, rinse each cartridge with 120 mL of reagent water,
                    without allowing the water to drop below the top edge of the packing.  If
                    the cartridge goes dry during the conditioning phase, the conditioning must
                    be started over.  Leave a small layer of reagent water on the top of the
                    cartridge. Begin adding the synthesized SUR solution to the cartridge.

            B.5.6.2  SAMPLE EXTRACTON - Adjust the vacuum so that the approximate
                    flow rate is 5 mL/min.  Do not allow the cartridge to go dry before all the
                    solution has passed through. After all of the  solution has passed through
                    the cartridge, rinse the cartridge with 50 mL of reagent water and draw air
                    or nitrogen through the cartridge for 5 min at high vacuum (10-15 in. Hg).

                    Note: If the reagent water rinse is not performed, substantial residual
                    sodium sulfite remaining in the cartridge will be eluted with the SUR.
                    Sodium sulfite is not soluble in methanol thus a white precipitate will be
                    observed. In this case, the extract will have to be taken to dryness,
                    reconstituted in reagent water and re-extracted according to Section B.5.6.

            B.5.6.3  CARTRIDGE ELUTION - Lift the extraction manifold top and insert a
                    rack with two 50-mL collection tubes into the extraction tank to collect the
                    extracts as they are eluted from the cartridge. Elute the analytes from the
                    cartridge with 50 mL of 10 mM ammonium acetate/methanol at a low
                    vacuum (~5 mL/min), such that the solvent exits the cartridge in a
                    dropwise fashion.

      B.5.7 EXTRACT CONCENTRATION - Concentrate the extracts under a gentle stream
            of nitrogen in heated water bath (60-65  °C) to 3-5 mL each. Weigh a 15-mL
            collection tube.  Transfer the two extracts to the weighed  15-mL collection tube
            rinsing with 2-3 mL of 10 mM ammonium acetate/methanol. Evaporate the extract
            to dryness.  Due to the amount of water (from the cartridges) in the extracts, the
            evaporation process will take in excess of 24 hours. After evaporation the clear
            residue will appear viscous.  Weigh the residue.

      B.5.8 Place the residue in a vacuum desiccator.  Periodically weigh the residue. Continue
            drying until constant weight is achieved. This may take several weeks.

B.6   VERIFICATION OF DIMETHACHLOR ESA

      B.6.1  Follow the procedure in Section 7.2.2 of the method to prepare a SUR PDS solution
            of the SUR near a concentration of 24 ug/mL in methanol.
                                        535-58

-------
B.6.2 Using the SUR PDS, prepare a "calibration standard" containing only the
      synthesized SUR at an appropriate concentration. During method development, the
      SUR concentration in standards and extracts was 120 ug/L.

B.6.3 Tune the LC/MS using the procedures outline in Section  10.2.1.1 and 10.2.1.2 of
      Method 535.

B.6.4 Initially, perform a full scan LC/MS run to locate the SUR peak and verify the
      molecular weight (dimethachlor ESA MW=301). Set the MS to scan at least
      200-400 daltons in the negative ion ESI mode.

B.6.5 Inject an aliquot of the "calibration standard" into the LC/MS using the suggested
      LC parameters outlined in Section 17, Table 1 of the Method 535.  Other LC
      conditions may be used at the discretion of the analyst. If other LC conditions are
      used, more emphasis will have be placed on the interpretation of the mass spectra
      since the SUR retention time will not match the retention time demonstrated in
      Method 535.

B.6.6 Locate a chromatographic peak containing an  [M-H]~ at m/z 300. Note the retention
      time.  In addition, carefully observe the whole chromatogram and verify that no
      other peaks (contaminants) are present which will interfere in the analysis of the
      other target analytes. Since the final analysis is MS/MS,  the main concern at this
      point should be whether any potential contaminants are present in large enough
      quantities to cause suppression in the electrospray source. A suggested guideline is,
      if the contaminant is present at lOx the SUR level, it may cause suppression of a
      co-eluting target analyte. If such a contaminant is present, further purification of the
      synthesized SUR maybe necessary. Figure Bl-A and Bl-B  demonstrate the
      relative purity of the synthesized standard. A  full scan mass spectra of synthesized
      dimethachlor  ESA is demonstrated in Figure Bl-C.

B.6.7 Optimize the LC/MS/MS using the procedures outlined in Section 10.2.1.3 of
      Method 535.  Confirm the identity of the synthesized SUR with LC/MS/MS by
      scanning all products ions produced by fragmentation of m/z 300.  Suggested
      MS/MS parameters are found in Tables 3 or 4 of the Method 535. Obtain a full
      scan MS/MS by scanning the product ions over a mass range of 75-305 daltons, if
      possible. (Ion trap instruments may not be able to scan this large of a mass range.
      In that case, use the lowest mass the ion trap will allow.)  A product ion mass
      spectra of synthesized dimethachlor ESA is demonstrated in Figure B2 using a triple
      quadrupole instrument. MS/MS fragmentation and relative abundances can vary
      significantly between instruments and depends heavily on various instrument
      settings. At a minimum, m/z 121  must be present in the synthesized SUR and if a
      triple quadrupole instrument is used, m/z 80 must be present. Other ions present in
      Figure B2 should be used, at the discretion of the analyst, to further verify the
      synthesis of the SUR.  Note:  m/z 212 was not observed in the ion trap product ion
      mass spectrum of dimethachlor ESA.

                                   535-59

-------
         loo                                    12.30
                  A) Full Scan LC/MS
                                     rtV*«^Wfr>i>&-^*^^
                                                 12.3O
      >-
     IS
                   B) m/z 300
                                           1O  rr^-    ,-  -  ^  15
                                              Time (mm)
                                                          3OO

                                                              [M-H]  of dimethachlor ESA
                C) TVIass spectrum of peak at 12.3O iiiin
                21 1
           200      220       240       2CO
                                                 280        300
                                                     m/z
                                                                    320       340        360       38O
Figure Bl.     LC/MS ananlysis of synthesized dimethachlor ESA demonstrating purity and confirmation of the molecular weight
              using a triple quadrupole mass spectrometer.
                                                           535-60

-------
                                                                        300

                                             212
                121
                12O    14O     16O    ISO    2OO
                                               22O   240
                                                    m/z
                                                           26O    28O    30O    320    34O    36O    38O    4OO
Figure B2.     Product ion mass spectrum of m/z 300 used for confirmation of synthesized dimethachlor ESA using a triple
              quadrupole mass spectrometer.
                                                         535-61

-------