EPA/600/R-00/066 September, 2000 INTEGRITY OF VGA-VIAL SEALS Brian A. Schumacher U.S. Environmental Protection Agency National Exposure Research Laboratory Characterization and Monitoring Branch P.O. Box 93478 Las Vegas, NV 89193-3478 and Martha M. Minnich, John H. Zimmerman, and J. Blasdell Lockheed Martin Environmental Services 980 Kelly Johnson Drive Las Vegas, Nevada 89119 Contract Number 68-C5-0091 NATIONAL EXPOSURE RESEARCH LABORATORY U.S. ENVIRONMENTAL PROTECTION AGENCY LAS VEGAS, NEVADA 89193-3478 ------- NOTICE The U.S. Environmental Protection Agency through its Office of Research and Development funded and managed the research described here under contract 68-C5-0091 to Lockheed-Martin Environmental Services. It has been subjected to the Agency's peer and administrative review and has been approved for publication as an EPA document. Mention of trade names or commercial products does not constitute endorsement or recommendation for use. -11- ------- Abstract Preservation of soil samples for the analysis of volatile organic compounds (VOCs) requires both the inhibition of VOC degradation and the restriction of vapor movement in or out of the sample container. The control of VOC vapor movement is generally assured through the use of screw caps with polytetrafiuoroethylene (PTFE) faced silicone septa that can be tightened to form a vapor lock within the volatile organic analysis (VGA) vial. The U.S. Environmental Protection Agency's (EPA) Region 7 laboratory expressed concern that visual imperfections in the glass lips and threads of VGA vials might allow VOCs to escape during storage. The objectives of this study were to determine if these imperfections lead to VOC losses and to identify an inexpensive screening test that could be used to distinguish between "defective" and competent vials. Clear, 40-mL glass VOA vials manufactured by the four major U.S. glass manufacturers were tested for seal integrity. The vials were purchased precleaned, with the manufacturer's or distributor's choice of PTFE faced silicone septa. All 216 vials from each manufacturer (864 vials, total) were visually inspected and then vacuum-tested for seal integrity by evacuating the vials, glass lips down against a polished aluminum plate. Visual inspection revealed a variety of imperfections ranging from small indentations, bumps, and scratches on vial threads or lips, through obvious defects, such as large indentations or grooves in the vial lips and chipped or broken glass. Imperfections were found on the lips or threads of 4% to 15% of the vials depending upon manufacturer. The aluminum plate vacuum test proved to be unreliable in identifying potentially leaky vials. No vials formed a complete seal regardless of the presence of visual imperfections. However, from each set of manufacturer's vials, the ten vials with the highest vacuum readings and the ten vials with the lowest vacuum readings (80 vials, total) while the pump was running were selected for two more tests, a septa-sealed vacuum test and a VOC-loss test. The septa-seal vacuum test was conducted twice on the 80 selected vials. No clear conclusion could be drawn about whether the flexibility of the silicone septa is sufficient enough to form a complete seal against VOC losses. For one manufacturer, the test was capable of identifying vials that would leak while for the other manufacturers, the test failed. The septa-seal vacuum test appears to be subject to a noticeable rate of false positives. Mean VOC concentrations after 14 days storage generally were within ± 20% of the known concentration with a majority of the concentrations within ± 15% of their known values. There were no statistically significant differences in VOC concentrations between vials in the potentially leaky and control group for any of the manufacturers. Only 1 vial lost VOCs and that was due to a large chip in the vial's lip and neck. These findings indicate that the silicone septa are flexible enough to overcome most vial imperfections and form a complete seal against VOC loss. A careful inspection of the VOA vials prior to use to remove any vials with large and obvious imperfections should be sufficient to screen out vials that are subject to VOC losses. -111- ------- Contents Page Notice ii Abstract iii List of Figures v List of Tables vi Acronyms vii Acknowledgments viii 1. Introduction 1 2. Materials 2 2.1 Vial Selection 2 2.2 Analytical Instrumentation 2 3. Procedures 3 3.1 Visual Inspection 3 3.2 Plate Vacuum Test 3 3.3 Septa-Sealed Vacuum Test 5 3.4 VOC Loss Test 5 3.5 VOC Gain Test 6 4. Results 6 4.1 Visual Inspection 6 4.2 Plate Vacuum Test 6 4.3 Septa-Sealed Vacuum Test 11 4.4 VOC Loss Test 14 4.5 VOC Gain Test 14 5. Summary and Conclusions 15 References 18 Appendix A - Visual Inspection Log and Plate Vacuum Test Results by Vial A-l Appendix B - Individual VOA Vial Analytical Results by Manufacturer B-l Appendix C - Quality Assurance/Quality Control Report for the Vacuum Studies C-l Appendix D - Quality Assurance/Quality Control Report for the VOC Loss Test D-l -IV- ------- List of Figures Figure 1. Aluminum plate vacuum test apparatus 4 -v- ------- List of Tables Table 1. Results of Plate Vacuum Test for Selected Vials 7 Table 2. Septa-Sealed Vacuum Test Results 12 Table 3. Summary of VOC-Loss Test Results 15 -VI- ------- Acronyms EPA United States Environmental Protection Agency HP Hewlett Packard PTFE polytetrafluoroethylene RSD relative standard deviation VOA volatile organic analysis VOC volatile organic compound -Vll- ------- Acknowledgments The authors would like to thank Mr. Don Miller of the EPA's Region 7 laboratory for his input and guidance throughout the initiation and running of this Regional Methods Initiative project. -Vlll- ------- Introduction Preservation of soil samples for the analysis of volatile organic compounds (VOCs) requires both the inhibition of VOC degradation and the restriction of vapor movement in or out of the sample container. VOC degradation is generally controlled by chilling the sample to 4° C (or in special cases, -20° C) and/or adding a preservative to the sample. For soils with VOC concentrations expected to be between 0.5 and 200 i-ig/kg, the addition of sodium bisulfate to reduce the sample's pH to < 2 is required (EPA, 1996). However, the sampler is cautioned that for soils containing carbonate minerals, the addition of sodium bisulfate is inappropriate due to sample effervescence effectively purging the sample prior to analysis. In contrast, for soils with high VOC concentrations (VOC concentrations > 200 |_ig/kg), the addition of methanol is required. Where the use of preservatives in the field is impractical or undesirable, a soil sample may be collected and temporarily stored in hermetically sealed samplers, such as the EnCore™ or SoilCore™ discrete samplers. Once the soil is sealed in the sampler, the sampler is chilled to at least 4° C for transport to the analytical laboratory. The collected soil sample is then transferred to a volatile organic analysis (VOA) vial and preserved (as appropriate) as soon as possible or, at least, analyzed within 48 hours of collection. The control of VOC vapor movement is generally assured through the use of polypropylene screw caps with polytetrafiuoroethylene (PTFE) faced silicone septa that can be tightened sufficiently enough to form a vapor lock within the VOA vial. Leaky VOA vial seals allow volatile contaminants to either escape or enter a sample; thereby, resulting in erroneous data. Sample integrity can be compromised by vapor losses or cross contamination that can occur during sample collection, transport, and storage. The current EPA method for the preparation of VOC samples, SW 846 Method 5035 (EPA, 1996), requires that VOA vials remain hermetically sealed until analysis by a purge-and-trap instrument capable of purging the sample by puncturing the septa of the open-top 40-mL VOA vial. If the samples remain hermetically sealed, the potential routes of sample vapor loss or contamination are limited to leaky seals or diffusion of VOCs through the septa. Poor vial sealing can occur due to: (a) the presence of sand grains (or other particles) on the lip or threads of the VOA vials; (b) septa expansion and contraction during temperature fluctuations associated with sample cooling/freezing or sample warming and cooling cycles during shipment from the field to the laboratory; or (c) puncturing the septa. The single-most important factor in sealing soil samples for VOC analysis is to insure that the lip and threads of the vial are clean before sealing. Traces of soil or grit on the glass lip will compromise the seals. Hewitt et al. (1995) intentionally left soil on the lip of three 135-mL (4 oz) bottles and thoroughly cleaned three others. Water (125 mL), spiked with a methanol solution containing four VOCs, was added to each of the six bottles and analyzed after 5 days storage at 4° C. Samples stored with dirty closures had an average of 41% to 46% lower VOC concentrations than the samples stored in the bottles where the threads and lips were cleaned prior to sealing. ------- Another potential cause of poor vial sealing is suspected to be slight aberrations in the lip and threads of the glass VOA vials. Quick preliminary visual examination of numerous VOA vials revealed that imperfections (e.g., indentations, grooves, and bumps) in the vial lips exist and may be large enough to provide a pathway for VOC loss. However, it is unclear whether the flexibility of the septa is sufficient enough to control the losses potentially associated with the imperfections in the VOA vials. Therefore, the study objectives were to: (1) identify the type and extent of imperfections in the lips or threads of VOA vials made by different glass manufacturers, (2) determine whether the imperfections lead to a loss of VOCs or if the septa creates a vapor tight VOA vial seal, and (3) determine if an inexpensive screening test could be used to distinguish between "defective" and competent vials. Materials and Methods Vial Selection Multiple VOA vial distributors (e.g. Fisher Scientific, Cole Farmer, I-Chem, Qorpak, Wheaton, etc.) were contacted to determine if they manufactured their vials or if they assembled the pieces purchased from different manufacturers. Most distributors indicated that they perform the following before marketing their products: (1) select a vial manufacturer and screen the glass vials to meet minimum quality control specifications, (2) select septa from any of a large number of laminators, (3) select a screw cap for the vial, and (4) perform precleaning steps, as necessary. As a result of the calls, it appears that four glass manufacturers supply nearly all the U.S. distributors with the 40-mL glass VOA vials. These manufacturers are: Chase Scientific Glass (Rockwood, TN), Comar Glass (Baltimore, MD), Kimble Glass Inc (Vineland, NJ), and Wheaton (Millville, NJ). The study was originally designed to include vials obtained directly from the major glass manufacturers. However, only one manufacturer offered precleaned vials. Therefore, the remaining three manufacturer's glass vials were ordered precleaned from distributors (i.e., Eagle- Pitcher, I-Chem, and Qorpak). A total of 864 vials (216 vials from each of the four major glass manufacturers) were examined. All vials were clear borosilicate glass and came with the manufacturer's or distributor's selection of open-top screw caps, and 3.05 mm silicone, 0.127 mm PTFE faced septa. Analytical Instrumentation VOCs were introduced by closed-system purge-and-trap (SW-846 Method 5035) into a gas chromatograph with mass spectrometer detector (MSB) following SW-846 Method 8260A except calibration was only for the analytes of interest (EPA, 1996). The Varian Archon™ purge-and-trap autosampler in conjunction with a Tekmar 3000 sample concentrator and a Vocarb 3000 trap was used to extract the samples. All samples were analyzed on a Hewlett- Packard (HP) Model 5890 Series II gas chromatograph with a HP Model 5970 mass spectrometer and a 60-m, 0.25-mm i.d. RTX Volatilization/RTX 5022 fused silica capillary column. The ------- MSB was scanned from 35 to 300 m/z at 70 eV in the El mode. The spectroscopic signal was analyzed using HP Chemstation software. Procedures Three types of tests, visual inspection, vacuum tests, and VOC measurements were used to assess the integrity of the vial seals. If a correlation between the visual or vacuum tests and the VOC measurements was observed, then one of the simple visual or vacuum tests may be offered as a screening method for a priori insurance that the vials to be used will seal properly. Visual Inspection All 216 vials from each manufacturer were uncapped and the glass lip and threads inspected for any defects. Slight imperfections consisted of: indentations or, conversely, bumps on the lip or threads; variability in the smoothness of the lip or threads; or variations in the glass thickness near the lip or threads. Obvious defects included chipped or broken glass, or clearly noticeable imperfections of the types described above. Vials were tracked by the serial number affixed to the vials, or numbered 1 through 216 if the vials did not have manufacturer assigned serial numbers (i.e., bar codes). Plate Vacuum Test A fiat plate vacuum test was conducted on all 216 vials from each of the manufacturers to determine if the vial lip imperfections prevented the formation of a vacuum tight seal. Vials were placed such that the vial lip was in direct contact with a polished aluminum plate (Fig. 1). In the center of the plate, a small hole was drilled and brass swagelok fittings were used to connect the open hole to a 0.25 horse-power vacuum/pressure pump (Gast Manufacturing Corporation, Benton Harbor, Ml). The swagelok nut on the top had the hex points ground nearly smooth to allow the 40-mL VOA vial (21.74 mm neck id) to fit directly over the nut. A grade AA accuracy vacuum gauge (Marsh PG-73, KW Instruments, Ontario CA) was connected between the vial and a valve used to isolate the gauge and vial from the pump. No grease or other sealant was used in this test. Vials were placed on the plate one at a time. The vacuum pump was turned on and a maximum vacuum reading was obtained in 30 seconds. While the vacuum pump was turned on, the readings remained stable. If the vial was isolated from the pump, the vacuum dissipated within 1 to 5 seconds indicating an incomplete seal between the aluminum plate and VOA vial. Therefore, the vacuum attained for each vial after 30 seconds was recorded for the test result. After conducting this test, data were ranked (lowest to highest vacuum obtained) for each set of vials. The 10 vials with the lowest vacuum readings from the plate vacuum test and 10 vials with the highest vacuum readings were selected as "potentially leaky" and control vials, respectively, for each manufacturer. These vials were designated by manufacturer (A through D) ------- Figure 1. Aluminum plate vacuum test apparatus. ------- and numbered 1 through 10 for the vials with the lowest vacuum readings and 20 through 29 for the vials with the highest vacuum readings. Septa-Sealed Vacuum Test In the second vacuum test, vials were evacuated with the septa and open-top screw caps fastened in place. The screw caps were hand tightened until a dimple formed in the center of the septum. Two needles were inserted into the septum, the first needle was connected to the vacuum gauge and the second needle was clamped to a hose connected to the vacuum pump. Sturdy 18-gauge needles were used in this test. Smaller gauge needles tended to core the septa and bend too easily, causing more problems than the heavy gauge needles. The silicone appeared to seal after puncturing, leaving a visible puncture mark in the PTFE only. The needles were inserted approximately 2 mm from and on opposite sides of the center point. Vials were evacuated to the capacity of the vacuum pump (i.e., 18620 ± 380 mm Hg). The evacuation needle was then removed leaving the vacuum gauge needle in place. The vacuum was noted and the needle was removed. After 3 days and 7 days, the vacuum remaining in each of the vials was measured. Puncture points for subsequent vacuum measurements were located approximately 2 mm off the center point and 90° from the prior insertion points. VOC Loss Test To monitor VOC losses from the sealed vials, VOC-spiked water was added to each of the 80 test vials. A bulk aqueous solution containing 50 ng/mL of each of the following analytes was prepared: acetone, benzene, chlorobenzene, 1,1-dichloroethene, methylene chloride, toluene, and trichloroethene. The spiked water was acidified with sulfuric acid to a pH < 2 and transferred to a Tedlar bag. Approximately 5 mL of VOC-spiked water was added to each of the test vials by gravity flow through PTFE tubing. Each vial was weighed before and after the addition of the spiked water to calculate the spike addition to individual vials. Vials were sealed, stored upright at 4° C, and VOC concentrations were measured after 14 days. Vials from the four manufacturers were prepared on four separate days to allow adequate time for the analysis. Duplicate samples were prepared by washing the vials after the first test was completed (water rinse, methanol rinse, and 105° C oven dry for 4 hours) and repeating the same spiking/storage/analysis procedure using a new septum for each vial. Internal standards (pentafiuorobenzene, 1,4-difiuorobenzene, chlorobenzene-d5, and 1,4-dichlorobenzene- d4) and system monitoring compounds (l,2-dichloroethane-d4, toluene-dg, and bromofiuorobenzene) were added by the Archon™ purge-and-trap autosampler just prior to analysis. ------- VOC Gain Test To monitor VOC gains from external contamination sources through the septa of sealed vials, 5 mL of deionized water were added to each of the 80 test VOA vials. The vials were capped with new, unpierced septa and placed in a 1 gallon paint can. An open vial containing 5 mL of methylene chloride was placed in the center of the paint can prior to closure. Samples were stored at room temperature for 14 days and then analyzed following SW-846 Methods 5035 and 8260A. Results Visual Inspection Imperfections were observed in 15 out of 216 vials from manufacturer A, 8 out of 216 vials from manufacturer B, 32 out of 216 from manufacturer C, and 13 out of 216 vials from manufacturer D (Appendix A). The imperfections noted for manufacturer A vials were minor indentations along the lip, which could not be felt while running one's finger around the lip, but could be seen when the vials were held up to the light and rotated. Two of the eight imperfect vials from manufacturer B were observed to have somewhat larger indentations on the lip and one vial had a chip in the glass threads. The remaining five vials from manufacturer B had minor indentations along the lip. Imperfections in the manufacturer C vials included: seven vials with dips (i.e., large indentations or grooves) in the glass or uneven lips (i.e., where one side of the lip was obviously lower than the other side); one vial with a chipped thread; one vial with a small bump on the lip; and the remainder of the vials had indentations that could be seen when held up to the light. One vial from manufacturer D had a major defect, a chip in the lip that extended down the side of the neck. One vial had a crack in its neck while minor chips in the neck were observed on six other manufacturer D vials. The remaining five vials had chips on their lips. Plate Vacuum Test The vacuum system was capable obtaining a vacuum of 18620 ±380 mm of Hg. Vacuum readings obtained in the vials after 30 seconds with the pump running ranged from 3800 to 19000 mm Hg (Appendix A). However, when the valve to the vacuum pump was turned off, the vacuum dissipated within 5 seconds. None of the vials sealed completely against the aluminum plate. Generally, the larger the visual imperfections, the worse the vial performed on the plate vacuum test. Very small or minor imperfections tended not to effect the results of the plate vacuum test. Vacuum readings for the "potentially leaky" vials ranged from 3800 to 13680 mm Hg (Table 1). In contrast, in the control vials, vacuum readings were consistently greater than 16720 mm Hg and were as high as a full vacuum reading of 18620 mm Hg. Interestingly, while no visible imperfections were identified for any of the control vials, both imperfect and visually ------- Table 1. Results of Plate Vacuum Test for Selected Vials. Vial ID Al A2 A3 A4 A5 A6 A7 A8 A9 A10 A20 A21 A22 A23 A24 A25 A26 A27 A28 A29 Serial No. 5979 6031 4523 5971 4530 6060 5929 6047 4564 6013 5988 6025 4515 4567 6057 4471 4536 6061 4563 4568 Visual defects Y Y Y Y Y Y Y N N Y N N N N N N N N N N Plate Vacuum (mm Hg) 5624 6840 7068 7220 7448 8208 8740 8740 9576 10336 18620 18620 18468 18468 18392 18316 18316 18316 18316 18240 ------- Table 1 . Results Vial ID Bl B2 B3 B4 B5 B6 B7 B8 B9 BIO B20 B21 B22 B23 B24 B25 B26 B27 B28 B29 of Plate Vacuum Serial No. 29200 29214 30547 29232 29163 29205 29228 29223 29189 30584 29166 30530 30568 29220 30630 29196 30586 29235 30525 30508 Test for Visual defects Y Y N N N Y N N N N N N N N N N N N N N Selected Vials (cont). Plate Vacuum (mm Hg) 5168 7144 10260 10640 10640 10868 11400 11476 11476 11628 18468 18468 18392 18240 18164 18164 17860 17860 17860 17860 ------- Table 1 . Results Vial ID Cl C2 C3 C4 C5 C6 C7 C8 C9 CIO C20 C21 C22 C23 C24 C25 C26 C27 C28 C29 of Plate Vacuum Serial No. 3291 5016 3982 4071 4351 4254 4399 3893 4583 3337 5025 4570 4143 4777 4360 3365 4490 3541 4623 4266 Test for Visual defects Y Y Y Y Y Y Y Y Y N N N N N N N N N N N Selected Vials (cont). Plate Vacuum (mm Hg) 3800 5320 5320 5548 5700 5776 6688 8208 9576 8576 17708 17480 17480 17480 17328 17328 17328 17328 17328 17328 ------- Table 1 . Results Vial ID Dl D2 D3 D4 D5 D6 D7 D8 D9 D10 D20 D21 D22 D23 D24 D25 D26 D27 D28 D29 of Plate Vacuum Serial No. 80 117 5 183 213 38 161 16 48 202 212 156 159 165 167 67 17 7 14 32 Test for Visual defects Y Y Y N N N N N N N N N N N N N N N N N Selected Vials (cont). Plate Vacuum (mm Hg) 5928 9728 12388 12388 12920 13148 13148 13680 13680 13680 17860 17708 17708 17480 17480 17480 17480 17480 17328 17328 10 ------- imperfection-free vials were found in the potentially leaky vials. In general, the imperfection- free vials in the potentially leaky group held the greater vacuum during the test. Septa-Sealed Vacuum Test The septa-sealed vacuum test was conducted twice to confirm the results. A loss of 3800 to 4560 mm Hg occurred each time the vacuum gauge was inserted into a vial caused by the dead air space in the vacuum gauge and needle. Vacuum readings that dropped by more than 4560 mm Hg by day 3 or more than 9120 mm Hg by day 7 were considered to be leaky. Vials that had readings of < 7600 mm Hg were considered to have catastrophically failed to seal. Vials that successfully held their vacuum after 3 and 7 days in at least one test were considered to be capable of forming a complete seal (i.e., the silicone septa was flexible enough to fill any potential leaks caused by the vial imperfections). Two vials from both the potentially leaky (vial numbers 1-10) and control groups (vial numbers 20 - 29) from manufacturer A were leaky after day 3 during the first test (Table 2). Three of those vials catastrophically failed to seal (i.e., vials A9, A21, and A24). After 7 days, 2 other vials, A5 and A28 had catastrophic seal failures. Upon retesting the vials, vial A24 again catastrophically failed and vials A21 and A28 successfully maintained its seal after 3 days but failed to hold a vacuum after 7 days. All the potentially leaky vials and 6 control vials made by manufacturer B were leaky, with one catastrophic leak in vial B3, during the first septa-sealed vacuum test (Table 2). Upon retesting, all of the potentially leaky vials and 3 of the originally identified leaky vials in the control group (i.e., vials B20, B21, and B28) remained leaky. Similar to the vials from manufacturer B, vials from manufacturer C gave very poor results on the first septa-sealed vacuum test with only one vial not being classified as leaky or having catastrophically failed after 3 or 7 days. It was noticed that the septa on these vials were more flexible than the septa on other vials. These septa would flex during puncturing by the needle gauge, causing leakage if the puncture was near the center. In the second test, all septa were punctured approximately 2-mm off the center point. During the retest, only 5 vials were deemed leaky and all those vials were from the potentially leaky group. The three vials from manufacturer D that had visual defects were leaky or failed catastrophically during the first septa-sealed vacuum test. During retesting, all three vials again either failed catastrophically (i.e., vials Dl and D3) or were leaky (i.e., vial D2) after 3 days and catastrophically failed after 7 days. Vial Dl had a major chip in the lip and was expected to leak. The glass lip of vial D3 was slightly raised on one side, but this vial had maintained a partial vacuum during the plate vacuum test, reaching a maximum vacuum of 12388 mm Hg (Table 1). Only one of the control group vials failed catastrophically during the first test; however, the same vial (i.e., D26) maintained its vacuum seal during the second septa-sealed vacuum test. 11 ------- Table 2. Septa-sealed vacuum test results Vacuum (mm Hg) Vial ID Al A2 A3 A4 A5 A6 A7 A8 A9 A10 A20 A21 A22 A23 A24 A25 A26 A27 A28 A29 Bl B2 B3 B4 B5 B6 B7 B8 B9 BIO B20 B21 B22 B23 B24 B25 B26 B27 B28 B29 Initial 18430 18430 18430 18430 18468 18468 18468 18620 18620 18620 18620 18620 18620 18620 18620 18544 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 Day 3 14250 14250 14250 14250 14250 13680 13680 11856 5016 14250 14592 6460 14653 14630 <1520 14630 14630 14440 14136 14630 12730 12730 4750 12350 11590 13718 10260 11438 13110 11552 13300 13110 14250 14136 14098 14098 13870 10602 8360 13718 Day 7 Initial-D 1 11590 11438 11400 11590 1900 10260 10070 7790 <1520 11210 11590 4560 11590 11590 <1520 11590 11590 11400 <1520 11590 8930 8930 <1520 8550 7258 10450 5700 7220 9500 7410 9880 9462 11020 11248 10868 11020 10488 5168 4180 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18240 18810 18810 18810 18810 19000 19000 19000 19000 19000 19076 19076 19152 19152 19190 19190 19190 19190 19190 18810 18810 18810 19000 19000 19000 19000 10450 19000 Day 3-D 14440 14440 14630 12540 14440 14630 14820 14440 14630 14440 14820 14630 15010 15010 <1520 14630 14820 14440 14440 6080 11590 11970 12730 2280 5320 14440 11970 11020 13490 13300 13490 8740 12350 8740 13300 13870 14440 14630 12350 14820 Day 7-D <1520 3230 11400 <1520 11020 11020 11400 11590 9880 11020 <1520 <1520 <1520 11590 <1520 10830 <1520 11020 9310 <1520 7030 7790 8550 <1520 1710 10830 7410 6460 9500 9120 9690 3800 9310 5510 9500 10450 11210 11780 7600 9120 12 ------- Table 2. Septa-sealed vacuum test results (cont.). Vacuum (mm Hg) Vial ID Cl C2 C3 C4 C5 C6 C7 C8 C9 CIO C20 C21 C22 C23 C24 C25 C26 C27 C28 C29 Dl D2 D3 D4 D5 D6 D7 D8 D9 D10 D20 D21 D22 D23 D24 D25 D26 D27 D28 D29 Initial 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 17480 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 18620 Day 3 <1520 <1520 12540 <1520 12008 <1520 <1520 14440 1710 <1520 <1520 9880 11210 2280 10450 <1520 14250 9348 14592 <1520 <1520 12388 <1520 14250 14250 14212 14250 14250 14250 14060 14288 14250 14250 14250 14212 14288 <1520 14250 14288 14288 Day 7 Initial-D 1 <1520 <1520 <1520 <1520 9690 <1520 <1520 4940 <1520 <1520 <1520 <1520 8550 <1520 8056 <1520 6080 4370 11590 <1520 <1520 8550 <1520 11210 11210 11020 11020 11058 11172 10830 11362 11020 11020 11210 11020 11248 <1520 11172 11210 19190 19000 19000 18620 19190 19190 19190 19190 19190 19190 19190 19190 19190 19190 19190 19190 19190 19228 19228 19228 18620 19380 19380 19380 19380 19380 19380 19380 19380 19380 19380 19380 19380 19380 19380 19380 19380 19380 19190 11210 19190 Day 3-D 14630 13110 8170 15010 15010 14820 12160 14630 13490 11400 14630 14820 15010 15010 15010 15010 15010 15010 15010 15010 <1520 11780 <1520 14630 14630 14630 14630 <1520 14630 14820 14630 14630 14630 <1520 14820 14630 14630 14630 14630 14630 Day 7-D 10830 8930 3420 11780 11780 11400 7600 10450 9500 6650 11020 11590 11780 11780 11780 11780 11780 11590 11780 11780 <1520 7220 <1520 11210 11210 11210 11210 <1520 11210 11210 11210 11210 11210 <1520 11020 11400 11210 11210 11020 11400 Second run for all vials labeled "-Dr 13 ------- VOC Loss Test Nearly all mean VOC concentrations after 14 days were within ± 20% of the initial VOC concentrations indicating little to no loss of VOCs during storage except for acetone in vials from manufacturer C (Table 3). The first analytical runs were performed on vials from manufacturer C. Acetone appeared to require a number of samples be run before the trap was properly "conditioned." Therefore, the "noise" in the acetone recoveries is unlikely to be related to the condition of the vial seals. The somewhat consistently lower VOC recoveries observed for vials from manufacturer C can be attributed to instrument instability (Table 3). Vials from each manufacturer were run as a batch, alternating the potentially leaky and control vials in sequence to even out instrument drift that tends to occur over the 14 hour batch runs (Appendices B and D). VOC concentrations in manufacturer C vials tended to be lower near the end of the analytical run, but recovered to the initial values before the run ended (Appendix B - Tables B2-5 and B2-6, Appendix D). The influence of this instrument fluctuation on VOC concentrations is difficult to assess; however, the lower concentrations are more likely due to an artifact of the analytical system rather than caused by leaky vial seals. Precision among the vials was acceptable with RSDs of < 20% with one exception (Table 3). A RSD of 26% was found for 1,1-dichloroethene in the control vials from manufacturer C. Upon examination of individual sample data, only vial Dl was clearly leaky (Appendix B). The chip in the side of the lip was so large that the vial would not seal. No data are reported for this vial because the recoveries of internal standards, added just prior to purging, were well below the QC criteria (Appendix D). No statistically significant differences were identified between VOC concentrations in the potentially leaky and control group vials from the same manufacturer (Table 3). The maximum absolute concentration difference between the two vial groups was 2 |j,g/mL. VOC Gain Test Only 3 vials from this test were analyzed due to overwhelming concentrations of methylene chloride found in each of the tested vials. The resultant chromatographic peaks were so broad and the concentrations so far above the calibration range that no further analytical testing was performed. The results of this test indicate that the septa could not prevent sample cross contamination by methylene chloride. Whether the cross contamination was due to poor seals or diffusion through the septa is not be positively known although the later explanation is most probable. 14 ------- Table 3. Summary of VOC-Loss Test Results. VOC Concentration, (ng/mL) A1-A10 Acetone 1 , 1 -Dichloroethene Methylene Chloride Benzene Trichloroethene Toluene Chlorobenzene Mean 36 40 57 44 38 40 38 RSD 8 5 2 5 5 5 5 A20-A29 Acetone 1 , 1 -Dichloroethene Methylene Chloride Benzene Trichloroethene Toluene Chlorobenzene Mean 37 40 57 44 38 39 37 RSD 5 5 4 5 5 5 5 B1-B10 Mean 37 40 61 41 38 37 37 RSD 11 8 5 7 8 8 8 B20-B29 Mean 37 41 61 42 38 37 37 RSD 11 7 3 5 8 8 8 C1-C10" Mean 28 35 53 40 33 35 34 RSD 7 17 6 10 9 9 9 C20-C29 Mean 27 34 53 39 32 34 33 RSD 7 26 9 15 19 14 9 Dl-D10f Mean 39 37 56 41 33 36 34 RSD 10 11 5 7 9 8 9 D20-D29f Mean 37 38 57 42 34 37 35 RSD 8 5 7 5 3 3 6 * - Mean and relative standard deviations (RSD) often potentially leaky vials (1 through 10) and ten control vials (20 through 29) for manufacturers A through D. ** - data are mean and RSDs of 8 vials (n=8). f - data are mean and RSDs of 9 vials (n=9). Summary and Conclusions Visual inspection of numerous VOA vials from the four major manufacturers in the United States found minor imperfections ranging from small indentations, bumps, scratches on vial threads or lips, through obvious defects, such as large indentations or grooves in the vial lips and chipped or broken glass. A less obvious imperfection that could affect the ability of a vial to 15 ------- seal completely was uneven rims in which one side of the vial's neck was clearly longer than the other side. Observed imperfections rates were 7, 4, 15, and 6% for manufacturer A, B, C, and D, respectively. Vials from manufacturer C had over twice as many imperfections (i.e., 32 out of 216) as vials from the other manufacturers. An aluminum plate vacuum test was developed as a means to perform a quick and simple test to determine if a vial will form a complete vacuum seal. Unfortunately, no matter how smooth the plate's surface was, none of the vials sealed completely against the plate. With the vacuum pump on, some vials were capable of reaching the pump's maximum pressure; however, when the pump was turned off, the vacuum quickly dissipated. These results make the value of the aluminum plate vacuum test in selecting whether or not a vial will obtain a complete seal against VOC loss highly questionable. A septa-sealed vacuum test was conducted to determine if the septa were flexible enough to form a complete seal even in the presence of vial imperfections. The results of this test were relatively inconsistent with the prior two tests. The vials from manufacturer A with visual imperfections did not lose their vacuum seal. In stark contrast, the only vials that failed to hold a vacuum were free from visual defects or were in the vial control group. All vials, with or without visual imperfections, in the potentially leaky group and 3 vials from the control group of manufacturer B were leaky. The aluminum plate vacuum test appeared to be a viable screening option for manufacturer B's vials although some false positives (i.e., vials falsely declared leaky) did occur. Vials made by manufacturer C had half of the vials in the potentially leaky group fail to hold their vacuum and the presence of a visual imperfection could not be used to clearly identify which ones. However, due to greater flexibility in the septa, the analyst had to be sure that the needle punctured the septa within 2 mm of the center or else the potential for seal leakage markedly increased. For manufacturer D, the presence of visual imperfections clearly indicated which vials would fail to hold their vacuum. All 3 vials with imperfections catastrophically failed to hold a vacuum while the remaining 7 vials identified as potentially leaky remained sealed after 7 days. No clear conclusion can be drawn based on the septa-sealed vacuum test about whether the flexibility of the silicone septa is sufficient enough to form a complete seal against VOC losses during storage. This test was subject to a noticeable rate of false positives. Mean VOC concentrations after 14 days storage generally were within ± 20% of the known concentration with a majority of the concentrations within ± 15% of their known values. There were no statistically significant differences in VOC concentrations between vials in the potentially leaky and control groups for any of the manufacturers. Only 1 vial lost VOCs and that was due to a large chip in the vial's lip and neck. These findings indicate that the silicone septa are flexible enough to overcome most vial imperfections and form a complete seal against VOC loss. Further, the use of the other tests (excluding visual inspection) performed in this study are unnecessary to screen for potentially leaky vials. A careful inspection of the VOA vials upon receipt (or at least, prior to use) to remove any vials with large and obvious imperfections 16 ------- should be sufficient to screen out vials that are subject to VOC losses through inadequate septum sealing. It should be noted that when puncturing the septa, the point of entry of the sparging needle on the purge-and-trap unit should be in the center (or within a few mm) of the septa. If it is not in or near the center, differences in septa flexibilities may result in a loss of the vial's vapor tight seal. A study to evaluate the properties of septa that influence their flexibilities and thus, their sealing potential, may be warranted. 17 ------- References EPA. 1996. Test methods for evaluating solid waste, SW-846. 3rd Edition. Office of Solid Waste and Emergency Response. U.S. Environmental Protection Agency, Washington, DC. December, 1996. Hewitt, A.D., T.F. Jenkins, and C.L. Grant. 1995. Collection, handling, and storage: Keys to improved data quality for volatile organic compounds in soil. Am. Environ. Lab. 2:25-28. 18 ------- APPENDIX A1 Visual Inspection Log and Plate Vacuum Test Results by Vial 1 NOTE: Vacuum readings presented in the following tables are directly from the vacuum gauge and have the units of inches Hg. To convert from inches Hg to mm Hg: 760 mm Hg = 1 inch Hg. ------- APPENDIX B Individual VOA Vial Analytical Results by Manufacturer ------- Table B2-1. Analytical Results for potentially leaky vials from manufacturer A . Al A2 A3 A4 AS A6 A7 A8 A9 A10 VOC Concentration, (ng/mL) Acetone 1,1-Dichloroethene Methylene Chloride Benzene Trichloroethene Toluene Chlorobenzene l,2-Dichlorethane-d4 Toluene-d8 Bromofluorobenzene 40 35 40 38 56 55 45 42 39 36 40 37 39 36 System 88 88 91 92 84 84 36 38 56 42 35 37 35 38 36 55 41 35 36 34 38D 39 57D 43 38 39 37 34 31 37 41 41 41 57 57 58 46 44 45 42 39 39 42 40 41 41 38 39 34 41 57 44 40 41 40 33 42 59 46 41 42 41 Monitoring Compound (SMC), % Recovery 90 93 83 89 93 83 87 91 83 - data superscripts indicate flagged data. Flag definitions are presented Table B2-2. Analytical Acetone 1,1-Dichloroethene Methylene Chloride Benzene Trichloroethene Toluene Chlorobenzene Results for control vials from A20 A21 A22 37 35 40 36 37 40 53 55 56 40 41 44 35 35 37 36 37 38 34 35 35 A23 42 39 56 43 36 37 35 manufacturer A24 VOC 35 39 56 42 36 37 35 A25 88 89 91 92 92 93 83 82 83 in Appendix D. A*. A26 A27 85 91 83 A28 85 90 83 A5D 26 36 41 39 32 33 38 76 73R 83 A29 A25D Concentration, (ng/mL) 38 43 58D 47 42D 42 40 System Monitoring Compound l,2-Dichloroethane-d4 Toluene-d8 Bromofluorobenzene 89 91 90 93 93 94 85 84 84 89 93 83 89 93 85 95 97 87 35 39 41 41 57 59 45 44 38 40 40 41 38 39 35 40 57 44 39 40 38 38 42 59 45 40 41 40 (SMC), % Recovery 89 91 93 93 83 84 89 92 82 91 92 81 29 36 38 38 30 32 34 78 74R 83 - data superscripts indicate flagged data. Flag definitions are presented in Appendix D. B- 1 ------- Table B2-3. Analytical Results for potentially leaky vials from manufacturer B*. Acetone 1,1-Dichloroethene Methylene Chloride Benzene Trichloroethene Toluene Chlorobenzene l,2-Dichlorethane-d4 Toluene-d8 Bromofluorobenzene Bl B2 39C 39C 43 37 61 58 44 40 41 37 40 37 40 37 System 89 87 89 90 87 85 * - data superscripts indicate flagged data. Flag Table B2-4. Analytical B3 VOC 37C 33 57 36 31 31 31 B4 B5 B6 B7 B8 B9 BIO Concentration, (ng/mL) 37C 38 60 40 37 36 36 40C 39C 35C 40 40 42 62 60 61 41 42 43 37 39 39 36 37 38 36 37 38 Monitoring Compound (SMC), % 89 90 84 89 91 84 88 88 88 90 91 91 85 83 84 definitions are presented in Appendix 27C 41C 39 61 40 35 34 34 42 63 41 38 37 38 37C 45 66 45 42 41 40 Recovery 87 90 84 D. 88 90 84 89 91 85 BSD 31C 50 58 46 37 39 36 94 89 82 Results for control vials from manufacturer B*. B20 B21 B22 B23 B24 B25 B26 B27 B28 B29 VOC Concentration, (ng/mL) Acetone 1,1-Dichloroethene Methylene Chloride Benzene Trichloroethene Toluene Chlorobenzene 34C 38C 39C 41 40 36 59 61 59 42 42 37 39 38 32 38 38 31 38 38 30 38C 42 63 43 40 38 38 39C 41 62 41 38 37 37 43C 37C 32C 40D 44 59 64 41 44 37 40 36 39 36 39 38 60 40 35 34 33 31C 43 63 44 41 39 40 36C 45 64 45 42 41 41 System Monitoring Compound (SMC), % Recovery l,2-Dichlorethane-d4 Toluene-d8 Bromofluorobenzene 87 88 87 89 89 89 88 87 85 84 90 83 88 90 85 88 89 90 91 84 86 87 90 84 87 90 89 86 91 85 B25D 38C 51 59 47 37 40 36 97 90 84 - data superscripts indicate flagged data. Flag definitions are presented in Appendix D. B-2 ------- Table B2-5. Analytical Results for potentially leaky vials from manufacturer C ' . Cl C2 C3 C4 C5 C6 C7 C8 C9 CIO VOC Concentration, (ng/mL) Acetone 1,1-Dichloroethene Methylene Chloride Benzene Trichloroethene Toluene Chlorobenzene l,2-Dichlorethane-d4 Toluene-d8 Bromofluorobenzene 29C 26C 40 33 53 52 43 38 37 32 39 34 38 33 System 90 91 89 89 80 81 - data superscripts indicate flagged data. Flag ** - nr = not reported (see text for discussion). Table B2-6. Analytical Acetone 1,1-Dichloroethene Methylene Chloride Benzene Trichloroethene Toluene Chlorobenzene 30C 38 56 41 34 36 34 25C 34 52 38 30 33 31 27C 27C 24C 28 40 6D 49 54 30D 34 43 17D 28 36 13D 31 38 17D 31 36 20D nrT nrT nrT nrT nrT nrT nrT 32C 27 50 35 29 32 32 26C 42 58 44 36 38 436 Monitoring Compound (SMC), % Recovery 89 88 81 88 89 83 89 87 87 89 89 90 81 80 79 nrT nrT nrT 93 90 82 90 89 79 C5D C7D C8D 27 32C 32C 46 50 48 56 58 57 40 46 44 32 37 33 34 39 37 30 35 33 88 90 89 94 95 98 91 86 88 definitions are presented in Appendix D. Results for control vials from manufacturer C*. C20 C21 C22 27C 29C 29C 39 33 39 55 53 56 43 37 40 36 30 33 38 33 35 36 32 33 C23 VOC 25C 38 55 41 34 36 34 C24 C25 C26 C27 C28 C29 Concentration, (ng/mL) 26C 42 58 44 38 38 37 29C 27C 43 21 56 43 44 30 38 24 39 27 37 28 25C 23 47 32 26 28 29 30C 22 47 32 25 29 29 26C 42 58 44 38 39 36 System Monitoring Compound (SMC), % Recovery l,2-Dichlorethane-d4 Toluene-d8 Bromofluorobenzene 91 89 90 89 89 89 89 82 82 89 90 83 87 88 79 90 86 91 89 82 80 87 89 82 91 90 80 89 89 80 C25D 32C 49 58 43 34 37 33 90 93 88 - data superscripts indicate flagged data. Flag definitions are presented in Appendix D. B-3 ------- Table B2-7. Analytical Results for potentially leaky vials from manufacturer D . Dl D2 D3 D4 D5 D6 D7 D8 D9 DIG VOC Concentration, (ng/mL) Acetone 1,1-Dichloroethene Methylene Chloride Benzene Trichloroethene Toluene Chlorobenzene l,2-Dichlorethane-d4 Toluene-d8 Bromofluorobenzene nr 37 nr 32 nr 51 nr 36 nr 26 nr 29 nr 26 System nd 90 nd 93 nd 84 46 29 52 36 29 33 32 38 37 56 41 33 36 35 35D 40 37 38 38 39 57 57 58 43 41 43 33 33 34 36 36 38 34 35 36 40 39 59 43 35 38 36 41 40 59 43 35 37 35 33 41 59 44 35 38 36 Monitoring Compound (SMC), % Recovery 92 91 84 * - data superscripts indicate flagged data. Flag definitions Table B2-8. Analytical Results for control vials from D20 D21 D22 D23 88 91 85 89 88 87 91 90 90 84 84 85 are presented in Appendix 86 90 83 D. 88 90 85 90 90 83 DSD 21C 38 50 41 29 34 31 87 85 84 manufacturer D . D24 D25 D26 D27 D28 D29 VOC Concentration, (ng/mL) Acetone 1,1-Dichloroethene Methylene Chloride Benzene Trichloroethene Toluene Chlorobenzene 35 39 36 34 36 38 50 54 56 39 41 42 31 33 34 34 36 38 32 34 35 44" nr 4" nr nr 2" 3" 33 38 56 42 33 37 34 40D 41 38 38 58 57 43 42 34 34 38 37 36 35 38 40 59 44 35 38 36 35 41 60 44 35 38 37 38 42 62 45 35 39 37 System Monitoring Compound (SMC), % Recovery l,2-Dichlorethane-d4 Toluene-d8 Bromofluorobenzene 90 89 90 92 91 92 83 84 83 92 85 84 88 91 83 85 89 90 91 84 85 91 91 84 87 90 85 87 89 84 D25D D25D 23C 27CR 38 39R 49 49R 40 41R 27 28R 33 34R 29 30R 89 92 84 86 82 83 - data superscripts indicate flagged data. Flag definitions are presented in Appendix D. - septum in vial PTFE-side up. Analytical results for information only, nr = not reported. B-4 ------- APPENDIX C Quality Assurance/Quality Control Report for the Vacuum Studies ------- The frequency and types of quality control samples for the vacuum studies followed or exceeded the specifications in the QAPP, "Integrity of VOA Vial Seals," January 7, 1998, Draft 1.2. The same vacuum pump and Grade AA vacuum gauge were used throughout the study. The maximum vacuum attained by the system was recorded each day before analyzing samples. Data for this calibration step are shown in Table A3-1. Table A3-1. Check for vacuum system integrity on days that vacuum readings were made. Date 12/10/97 12/15/97 12/21/97 12/22/97 12/23/97 3/31/98 4/3/98 4/6/98 4/7/98 4/10/98 4/13/98 Vacuum (mm Hg) 18620 18620 18620 18620 17860 18088 18468 18620 18620 18620 19380 Duplicate vacuum readings were performed every 10 samples for the plate-vacuum study and for all samples in the septa-seal vacuum study. Data for the duplicate samples are given in Tables B2-1 through B2-8. A summary of the percent difference (%D) for the plate-vacuum study duplicates is given in Table A3-2. The reproducibility of the test was less than anticipated with RPD values ranging from zero to 13.5%. The mean RPD was 3.8%. C-l ------- Table A3-2. Relative percent differences for duplicate vacuum plate samples. Data presented are for 22 duplicate samples out of 216 vials from each manufacturer. Manufacturer Manufacturer Manufacturer Manufacturer A B C D 6.3 12.5 6.5 3.6 5.7 1.6 3.1 5.1 3.1 1.8 3.0 6.2 1.1 1.8 5.5 1.1 4.5 1.1 1.1 0.8 0.6 1.0 3.6 0.7 6.7 3.8 3.6 5.4 10.4 11.5 9.1 6.2 0.9 4.3 3.4 7.7 0 8.2 6.3 3.4 4.5 8.7 2.6 1.1 10.9 0 8.7 0 4.8 13.5 0.9 9.5 6.1 7.9 1.3 0 2.5 8.2 11.5 0.5 6.3 2.3 2.8 1.1 1.1 1.1 12.1 2.7 0 1.3 2.5 0 1.2 1.2 1.2 1.2 1.2 0 1.1 8.4 0.5 1.1 1.8 3.5 0 7.0 0 2.2 C-2 ------- APPENDIX D Quality Assurance/Quality Control Report for the VOC Loss Test ------- The types of quality control samples, acceptance criteria, and qualifiers given in Table A4-1 reflect those listed in Table 1 of the QAPP "Integrity of VGA Vial Seals," January 7, 1998, Draft 1.2. These qualifiers appear in Tables B2-1 through B2-8, where appropriate. TABLE A4-1. VOC PC SAMPLES and ACCEPTANCE CRITERIA QC Sample Duplicates Continuing calibration Internal standards SMC recovery IDL (7 replicates) Acceptance Criteria % D < 25% for all target analytes % D < 25% for all target analytes Area counts 50-200% of 50 [ig/mL std. % R = 100 ± 25% for all target analytes % R = 100 ± 25% for SMCs Qualifiers flag "D" flag "C" flag "T" flag "R" -- Duplicate Samples One duplicate sample for each set often "potentially leaky" or "control" vials was reported. Additional duplicate data are given for samples C7, C8, and D23 because the initial data were questionable (either the internal standard or the system monitoring compound recovery was poor). Instrument Detection Limits The instrument detection limits (IDLs) were determined as 3.143 times the standard deviation obtained from the analysis of 7 replicates of the calibration standard at a nominal concentration of 5 ng/mL. The system monitoring compound (SMC) recoveries for all compounds were within the ± 25% as specified. By this method, the detection limits, in ng/mL, for the compounds were: acetone = 3.5, 1,1-dichloroethene = 0.8, methylene chloride = 1.0, benzene = 0.3, trichloroethane = 0.3, toluene = 0.7, and chlorobenzene = 0.5 ng/mL. These detection limits are approximately 2 orders of magnitude lower than the 50 ng/mL working VOC concentrations required for this work. Instrument Blanks Instrument blanks were run prior to the analysis of an initial 10 ng/mL standard used in a 5 point initial calibration curve. Blanks were not run following the CCVs (continuing calibration verification) prior to sample analysis. During preliminary analysis of VOC samples, it became evident that the first sample of a batch run immediately following a blank yielded low concentrations for the target analytes relative to the remainder of the batch. This was attributed to the blank which conditioned the trap differently than a sample. Because the CCV closely matches the nominal 50 ng/mL concentration of the sample target analytes, analysis of the CCV D- 1 ------- conditions the trap in the purge and as if it were a sample. Since the objective of this project was to determine differences in VOC concentrations as a consequence of vial seal integrity, not trap dependency of previously run samples, analysis of blanks following a CCV was eliminated. Continuing Calibration Continuing calibration data are shown in Tables D2-1 through D2-4. Analytes in the CCVs that did not meet the 25% D window are flagged. Samples that were run immediately prior to the CCV that did not meet the 25% D window are also flagged. The polar nature of acetone yielded widely variable recoveries for this analyte. Acetone failed the CCV most often while the remaining analytes easily met criteria. It was deemed impractical to run new calibration curve solely for acetone. This forced some portions of an analytical batch to be run with acetone failing the midpoint CCV. These samples are flagged. SMC Recovery The system monitoring compounds, l,2-dichloroethane-d4, toluene-dg, bromofluorobenzene were added to every sample and CCV. SMC recoveries of samples are reported with the sample data (Tables B2-1 through B2-8). Those samples not meeting the ± 25% R criteria are flagged. Samples Analytical data for target analytes are presented in Tables B2-1 through B2-8. Each batch of 20 samples from one vendor were run in the sequence of CCV1 XI, X20, X2, X21, X3, X22, X4, X23, X5, X24, CCV2, X6, X25, X7, X26, X8, X27, X9, X28, X10, X29, and CCV3 where X represents the vendor letter identification. Tables B2-1 and B2-2 contain VOC test results from Manufacturer A vials. For vial A5, acetone and methylene chloride did not meet %D criteria in the sample duplicate and were flagged. Variability in duplicates was not unexpected. The duplicate samples came from 2 different analytical runs using two different spiking solutions that were quantified from two different initial calibration curves. The polar nature of acetone introduces further variations in recoveries depending on the conditioning of the trap as has been discussed previously. In the sample duplicate A5D, recovery of the SMC, toluene-d8, was 2% low and was flagged. The remaining 2 SMCs were low, but passed. For vial A25 in Table B2-2 methylene chloride and trichloroethene did not meet %D criteria in the sample duplicate and were flagged. In the sample duplicate A25D, recovery of the SMC, toluene-dg, was low and was flagged. The remaining 2 SMCs were low but passed. D-2 ------- Tables B2-3 and B2-4 contain VOC test results from Manufacturer B vials. The target analyte acetone failed high in CCV2. Because this was a mid batch CCV, the samples preceding the CCV which include, Bl, B20, B2, B21, B3, B22, B4, B23, B5, and B24 and the samples following the CCV which include B6, B25, B7, B26, B8, B27, B9, B28, BIO, and B29 were flagged for acetone. The target analyte acetone failed high in CCV2D. Because this was a mid batch CCV, the sample BSD preceding the CCV and the sample B25D following the CCV were flagged for acetone. For vial B25 in Table B2-4, 1,1-dichloroethene did not meet %D criteria in the sample duplicate and was flagged. Tables B2-5 and B2-6 contain VOC test results from Manufacturer C vials. The target analyte acetone failed low in CCV2 and CCV3. Consequently samples Cl, C20, C2, C21, C3, C22, C4, C23, C5, C24, C6, C25, C7, C26, C8, C27, C9, C28, CIO, and C29 were flagged for acetone. The target analyte acetone also failed low in CCV3D. Because this was a closing CCV, the sample duplicates C7D and C8D preceding this CCV were flagged. Vial C7 in Table B2-5 did not meet the %D criteria for all analytes and was flagged although the SMCs met their QC criteria. Visual inspection of the vial did not reveal observable defects. It is suggested the low recoveries of the analytes are a consequence of a loose cap due to incomplete tightening rather than defects in the sealing lip of the vial. Vial C8 in Table B2-5 did not meet the %D criteria for all analytes and the recoveries for the SMCs were all significantly below 50% so this sample was flagged. Visual inspection of the vial did not reveal observable defects. It is suggested as for vial C7, the loss of analytes and SMCs are a consequence of a loose cap causing VOCs to be lost during the purge cycle rather than defects in the sealing lip of the vial. Tables B2-7 and B2-8 contain VOC test results from Manufacturer D vials. The target analyte acetone failed low in CCV2D. Consequently samples DSD, D23D, and D25D were flagged for acetone. The CCV and the sample D25D following the CCV were flagged for acetone. The recovery for the SMC 1,2-dichloroethane-d4 was low in CV3D and all analytes in sample D25D were "R" flagged. Analytical results for sample Dl in Table B2-7 was not reported. The glass lip of the vial was so badly chipped, that a seal could not be obtained between it and the septum cap. Any remaining VOCs as well as SMCs added immediately prior to the purge were lost. Sample D5 did not meet %D criteria for acetone and was flagged. The septum cap in vial D23 had the Teflon-septum facing exterior to the vial which allowed permeation of VOCs through the septum. Interestingly, the polar VOC acetone, showed no loss by permeation into/through the silicone septum. All SMCs in the sample meet criteria indicative of a valid purge. Analytical D-3 ------- data for sample D23 are for information only as the vial cap was improperly configured. Sample D25 did not meet %D criteria for acetone and was flagged. D-4 ------- Table D2-1. Continuing calibration results for samples and duplicate samples from manufacturer A. CV1 CV2 CV3 CV1D CV2D CV3D VOC Concentration, (ng/mL) Acetone 1,1-Dichloroethene Methylene Chloride Benzene Trichloroethene Toluene Chlorobenzene 46 52 58 52 52 51 52 50 48 57 48 50 49 50 57 49 55 47 49 48 50 System Monitoring l,2-Dichlorethane-d4 Toluene-d8 Bromofluorobenzene Table D2-2. Continuing 97 103 104 calibration CV1 98 99 100 95 96 98 results for samples CV2 CV3 57 43 41 45 42 43 50 Compound (SMC), % 86 82 98 48 41 43 44 42 44 51 40 40 42 45 42 44 50 Recovery 90 83 102 and duplicate samples from CV1D CV2D 188 83 98 manufacturer A CV3D VOC Concentration, (ng/mL) Acetone 1,1-Dichloroethene Methylene Chloride Benzene Trichloroethene Toluene Chlorobenzene 59 50 52 48 51 48 51 64C 51 54 49 52 50 51 40 46 50 46 49 47 49 System Monitoring l,2-Dichlorethane-d4 Toluene-d8 Bromofluorobenzene 95 98 106 99 102 105 93 97 102 45 54 52 51 48 50 48 Compound (SMC), % 101 100 97 71C 57 60 53 52 53 51 43 53 57 53 48 50 48 Recovery 113 106 105 110 101 100 - data superscripts indicate flagged data. Flag definitions are presented in Table A4-1. D-5 ------- Table D2-3. Continuing calibration results for samples and duplicate samples from manufacturer C CV1 CV2 CV3 CV1D CV2D CV3D VOC Concentration, (ng/mL) Acetone 1,1-Dichloroethene Methylene Chloride Benzene Trichloroethene Toluene Chlorobenzene 43 51 51 52 51 51 52 36C 31C 48 48 50 50 49 50 49 49 49 50 50 50 System Monitoring l,2-Dichlorethane-d4 Toluene-d8 Bromofluorobenzene - data superscripts indicate 104 101 103 flagged data. 98 99 99 101 101 102 Flag definitions Table D2-4. Continuing calibration results for samples CV1 CV2 CV3 50 56 57 53 53 55 53 Compound (SMC), % 105 109 110 are presented in Table 40 54 56 51 51 47 50 37C 57 60 54 51 54 49 Recovery 99 102 105 A4-1. and duplicate samples from CV1D CV2D 111 112 102 manufacturer D CV3D VOC Concentration, (ng/mL) Acetone 1,1-Dichloroethene Methylene Chloride Benzene Trichloroethene Toluene Chlorobenzene 50 48 52 49 51 50 51 47 52 49 48 53 53 57 48 51 51 49 49 51 52 System Monitoring l,2-Dichlorethane-d4 Toluene-d8 Bromofluorobenzene 99 100 104 96 98 100 101 104 114 50 53 52 52 52 52 51 35C 44 51 47 46 47 47 44 49 45 42 41 41 39 Compound (SMC), % Recovery 107 104 102 97 95 99 56R 81 79 - data superscripts indicate flagged data. Flag definitions are presented in Table A4-1. D-6 ------- |